vvEPA
         United States
         Environmental Protection
         Agency
           Office of Research and
           Development
           Washington DC 20460
EPA/600/R-95/131
August 1995
Methods for the
Determination of
Organic Compounds in
Drinking Water
         Supplement

-------

-------
                                       EPA-600/R-95/131
                                           AUGUST 1995
   METHODS FOR THE DETERMINATION

        OF  ORGANIC COMPOUNDS

         IN DRINKING WATER
           SUPPLEMENT III
      ERRATA
NATIONAL EXPOSURE RESEARCH LABORATORY
 OFFICE OF RESEARCH AND DEVELOPMENT
U.S. ENVIRONMENTAL PROTECTION AGENCY
       CINCINNATI, OHIO  45268
                                       Printed on Recycled Paper

-------
                                  DISCLAIMER
     This manual has been reviewed by the National Exposure Research
constitute endorsement or recommendation for use.

-------
                          ERRATA
Methods for the Determination of Organic Compounds in Drinking Water
                Supplement III, EPA/600-R-95/131

-------

-------
iRRATA - Nov. 27, L995
There is an error in the sample size in Section 8.1.   This sheet corrects that
mges 515.2-11 and 515.2-12 and replace them with this sheet.     corrects that



                             they are  certified  by the manufacturer or by an
                             independent source.

                   7.17.2    Transfer  the stock  standard  solutions .into  15-ml
                             TFE-fluorocarbon-sealed  screw cap  amber vials.  Store at
                             4°C or less when  not in  use.

                   7.17.3    Stock standard solutions should be replaced after 2 months
                             or sooner if comparison with laboratory fortified blanks
                             or QC samples indicate a problem.

                   7.17.4    Primary Dilution Standards — Prepare two sets of
                             standards according to the sets labeled A and B in Table
                             1.  For each set, add approximately 25 ml of methanol  to a
                             50 ml volumetric flask.   Add aliquots of each stock
                             standard in the range of approximately 20 to 400 uL and
                             dilute to volume with methanol.   Individual  analyte
                             concentrations will  then be in the range of 0.4 to 8 ug/mL
                             (for a 1.0 mg/mL stock).  The minimum concentration would
                             be appropriate for an analyte with strong electron capture
                             detector (ECD)  response, e.g. pentachlorophenol.   The
                             maximum concentration is for an  analyte with weak
                             response,  e.g.,  2,4-DB.   The concentrations  given  in Table
                             2  reflect  the  relative  volumes of stock standards  used for
                             the primary dilution standards used in generating  the
                             method  validation data.   Use these relative  values to
                             determine  the  aliquot volumes of  individual  stock  stan-
                             dards  above.

              7.18  INTERNAL STANDARD  SOLUTION  -  Prepare  a stock internal  standard
                   solution by  accurately weighing approximately 0.050 g  of  pure  DBOB
                   Dissolve the DBOB  in methanol  and  dilute to  volume  in  a 10-mL
                   volumetric flask.  Transfer the DBOB solution to  a TFE-fluorocarbon-
                   sealed screw cap bottle and store  at room temperature.  Prepare  a
                   primary  dilution standard at approximately 1.00 /ig/mL  by  the addi-
                   i°?nnf  ?° 2L*uf tne^ock  standard to  100 mL of methanol.  Addition
                   of 100 nL of the primary dilution  standard solution to the final 5
                  mL of sample extract (Sect.  11) results 1n a  final internal standard
                   concentration of 0.020 pg/ml.   Solution should be replaced when
                  ongoing  QC (Sect. 9) Indicates  a problem.  Note that DBOB has been
                   shown to be  an effective Internal standard for the method analytes,
                  but other compounds may be  used if the QC requirements in Sect  9
                  are met.

             7.19 SURROGATE ANALYTE SOLUTION - Prepare a surrogate analyte stock
                  standard solution by accurately weighing approximately 0.050 g of
                  pure DCAA.  Dissolve the DCAA in methanol and dilute to volume in a
                  10-mL volumetric flask.  Transfer the surrogate analyte solution to
                  a TFE-fluorocarbon-sealed screw cap bottle and store at room temper-
                  ature.  Prepare a primary dilution standard at approximately 2.0
                  Mg/mL by addition of 40 pL at the stock standard to 100 mL of
                  methanol.  Addition of 250 /tL of the surrogate analyte  solution to a
                  250-mL sample prior to extraction  results 1n a surrogate concentra-
                                           515.2-11
                                                             Rev.  1.1,  Aug.  1995

-------
          tion in the sample of 2 /tg/L and, assuming quantitative recovery of
          DCAA, a surrogate analyte concentration in the final 5 ml extract of
          0.1 /ig/mL.  The surrogate standard solution should be replaced when
          ongoing QC (Sect. 9) indicates a problem.  DCAA has been shown to be
          an effective surrogate standard for the method analytes, but other
          compounds may be used if the QC requirements in Sect. 9 are met.

     7.20 INSTRUMENT PERFORMANCE CHECK SOLUTION — Prepare a diluted dinoseb
          solution by adding 10 /iL of the 1.0 fig/pi dinoseb stock solution to
          the MTBE and diluting to volume in a 10-mL volumetric flask.  To
          prepare the check solution, add 40 ^L of the diluted dinoseb solu-
          tion, 16 nl of the 4-nitrophenol stock solution, 6 0L of the 3,5-
          dichlorobenzoic acid stock solution, 50 (il of the surrogate standard
          solution, 25 pi of the internal standard solution, and 250 pi of
          methanol to a 5-mL volumetric flask and dilute to volume with MTBE.
          Methylate sample as described in Sect. 11.4.  Dilute the sample to
          10 ml in MTBE.  Transfer to a TFE-fluorocarbon-sealed screw cap
          bottle and store at room temperature.  Solution should be replaced
          when ongoing QC (Sect. 9) indicates a problem.

8.   SAMPLE COLLECTION. PRESERVATION. AND STORAGE

     8.1  Grab samples should be collected in 250 ml amber glass containers.
          Conventional sampling practices (7) should be followed; however, the
          bottle must not be prerinsed with sample before collection.

     8.2  SAMPLE PRESERVATION ANQ STORAGE

          8.2.1  If residual chlorine Is present, add 80 mg of sodium thiosul-
                 fate (or 50 mg of sodium sulfite) per liter of sample to the
                 sample bottle prior to collecting the sample.  Demonstration
                 data in Section 17 of this method was obtained using sodium
                 thiosulfate.              .r.-_

          8.2.2  After the sample is collected in.the bottle containing the
                 dechlorinating agent, seal the bottle and mix to dissolve the
                 thiosulfate.    ,        .  ...
                               /*.-        '  '
          8.2.3  Add hydrochloric add (diluted 1:1 in reagent water) to the
                 sample at the sampling site in amounts to produce a sample pH
                 £ 2.  Short range (0-3) pH paper (Sect. 6.14) may be used to
                 monitor the pH.  Note: Do not attempt to mix sodium thiosul-
                 fate and HC1 in the sample bottle prior to sample collection.

          8.2.4  The samples must be iced or refrigerated at 4"C away from
                 light from the time of collection until extraction.  Preser-
                 vation study results Indicate that the sample analytes (mea-
                 sured as total add), except 5-hydroxy-dicamba, are stable in
                 water for 14 days when stored under these conditions (Tables
                 8 and 9).  The concentration of 5-hydroxydicamba is seriously
                 degraded over 14 days in a biologically active matrix. How-
                 ever, analyte stability win very likely be affected by the
                                   515.2-12

-------
*
  ERRATA;  Nov.  27,  1995
                            7.19) to each  250-mL- sample.  The  surrogate will be  at a
                            concentration  of  2 ng/l.  Dissolve  50 g sodium sulfate in the
                            sample.      .  ,.         ,   ;   ..... t>;

                     11.1.3 Add 4 ml of 6  N NaOH  to each  sample, seal, and shake.  Check
                            the pH of the  sample  with pH  paper or a pH meter;  if the
                            sample does not have  a pH greater  than or equal to 12, adjust
                            the pH by adding  more 6 N NaOH.  Let the sample sit  at room
                            temperature for 1 hr, shaking the  separatory funnel  and
                            contents periodically.  Note: Since many of the herbicides
                            contained in this method are  applied as a variety of esters
                            and salts,  it  is  vital to hydrolyze them to the parent acid
                            prior to extraction.  This step must be included in  the
                            ana^ls of a11 extracted field samples, LRBs, LFBs, LFMs
                            and QCS.                                                 '

                     11.1.4 Use 15 mL mmethylene  chloride to rinse the sample bottle and
                            the granduated cylinder.  Then transfer the methylene
                            chloride to the separatory funnel and extract the sample bv
                            vigorously  shaking the funnel for 2 min with periodic ventinq
                            to release  excess pressure.   Allow the organic layer to  sepa-
                            rate from the water phase for a minimum of 10 min.   If the
                            emulsion interface between layers is more than one-third the
                            volume of the solvent layer,  the analyst must employ
                            mechanical  techniques to complete the phase separation   The
                         .   optimum technique depends  upon the sample,  put may  include
                            stirring, filtration through  glass wool,  centrifugation,  or
                            other Physical  methods.   Discard the methylene chloride  phase
                            (Sect. 14, 15).

                     11.1.5 Add a second  15-ml volume  of  methylene  chloride  to  the separ-
                            atory funnel  and  repeat  the extraction procedure  a  second
                            time,  discarding  the  methylene chloride  layer.   Perform a
                            third  extraction  in  the  same  manner.

                     11.1.6 Drain  the contents of the.  separatory  funnel  into a  500-mL
                            beaker.  Adjust the pH to  1.0 ±  0.1 by the dropwise addition
                            of concentrated sulfuric acid with  constant stirring. Monitor
                            the pH with a pH meter (Sect.  6.8)  or short range (0-3) oH
                            paper  (Sect. 6.14).                                   ' v

               11.2  SAMPLE EXTRACTION

                     11.2.1  Vacuum Manifold — Assemble a  manifold (Sect. 6.3) consistinq
                            °f H*3?"11111 f1asks Wlth f^ter funnels (Sect. 6.1,6.2).
                            Individual vacuum  control, on-off and vacuum release  valves
                            and vacuum gauges  are  desirable.  Place the 47 mm extraction
                           disks (Sect. 7.1)  on the filter frits.

                     11.2.2 Add 20 mL of 10% by volume of methanol In MTBE to the top of
                           each disk without  vacuum and allow the solvent to remain for
                           2 min.  Turn on full vacuum and draw the solvent through the
                           disks, followed by room air for 5 min.

                                             515.2-19


                                                                   Rev.  1.1,  August  1995

-------
     11.2.3 Adjust the vacuum to approximately 5 in.  (mercury)  and  add
            the following in series to the filter funnel  (a)  20 ml
            methanol  (b)  20 ml reagent water (c) sample.   Do  not allow
            the disk  to dry between steps and maintain the vacuum at  5
            in.

     11.2.4 After all  the sample has passed through the disk,  apply
            maximum vacuum and draw room air through the disks  for  20
            min.

     11.2.5 Place the culture tubes (Sect. 6.4) in the vacuum tubes to
            collect the eluates.  Elute the disks with two each 2-mL
            aliquots  of 10% methanol in MTBE.  Allow each aliquot to
            remain on the disk for one min before applying vacuum.

     11.2.6 Rinse each 500-mL beaker (Sect. 11. 1.6) with 4 ml  of pure  MTBE
            and elute the disk with this solvent as in Sect.  11.2.5.

     11.2.7 Remove the culture tubes and cap.

11.3 EXTRACT PREPARATION

     11.3.1 Pre-rinse the drying tubes (Sect. 7.5.1)  with 2 ml  of MTBE.

     11.3.2 Remove the entire extract with a 5-mL pipet and drain the
            lower aqueous layer back Into the culture tube.  Add the
            organic layer to the sodium sulfate drying tube (Sect.
            7.5.1).  Maintain liquid 1n the drying tube between this  and
            subsequent steps.  Collect. the dried extract in a 15-mL
            graduated centrifuge tube or a 10-mL Kuderna-Danish tube.

     11.3.3 Rinse the culture tube with an additional 1 mL of MTBE  and
            repeat Sect.  11.3.2.      •-.-

     11.3.4 Repeat step Sect. 11.3.3 and finally add a 1-mL aliquot of
            MTBE to the drying tube before it empties.  The final volume.
            should be 6-9 ml.  In this form the extract is esterified as
            described
11.4 EXTRACT ESTERIFICATION WITH DIAZOMETHANE ~ See Section 11.5 for
     alternative procedure.

     11.4.1 Assemble the dlazomethane generator (Figure 1) in a hood.

     11.4.2 Add 5 ml of ethyl ether to Tube 1.  Add 4 ml of Oiazald
            solution (Sect. 7.12) and 3 ml of 37X KOH solution (Sect.
            7.16.1) to the reaction tube 2.  Immediately place the exit
            tube Into the collection tube containing the sample extract.
            Apply nitrogen flow (10 mL/m1n) to bubble dlazo-methane
            through the extract.  Each charge of the generator should be
            sufficient to ester ify four samples.  The appearance of a
            persistent yellow color 1s an Indication that esterifi cat ion
            1s complete.  The first sample should require 30 sec to 1 min

                              515.2-20

-------
      - Nov. 27,1995            """'•"'' .......
here is an equation missing in Sect. 10.4.3.  This sheet corrects that error   Remove
24.2*21 and 524. -22 and replace them with this sheet.



                   10.3.4 Determine that the absolute areas of the quantitation ions of
                          the internal  standard and .surrogates have not decreased by
                          more than 30% from the areas measured in the most recent
                          continuing calibration check, or by more than 50% from the
                          areas measured during initial calibration.   If these areas
                          have decreased by more than these amounts,  adjustments must
                          be made  to restore system sensitivity.   These adjustments may
                          require  cleaning of the MS ion source,  or other maintenance
                          as indicated  in Sect. 10.3.6, and recalibration.  Control
                          charts are useful  aids in documenting system sensitivity
                          changes.                                                J

                   10.3.5 Calculate the RF for each analyte of concern and surrogate
                          compound  from the  data measured  in the  continuing calibration
                          check.  The RF for each analyte  and surrogate must be within
                          30%  of the mean value measured in the initial  calibration.
                          Alternatively,  if  a linear or second order  regression is
                          used,  the concentration measured  using  the  calibration curve
                          must  be within  30% of the true value of the concentration in
                          the  calibration solution.  If these conditions  do not exist
                          remedial  action must  be taken which may require recalibrati-
                          on.   All  data from field samples  obtained after the  last
                          successful  calibration  check  standard,  should  be considered
                          suspect.  After remedial  action has been taken,  duplicate
                          samples should  be  analyzed  if they are  available.

                  10.3.6  Some  possible remedial  actions.   Major  maintenance such  as
                          cleaning  an ion  source,  cleaning  quadrupole  rods, etc   re-
                          quire returning  to the  initial calibration  step.

                          10.3.6.1  Check  and adjust,.GC  and/or MS  operating  conditions-
                                   check the MS resolution, and calibrate  the mass
                                   scale.

                          10.3.6.2 Clean 'or  replace, the splitless injection liner;
                                  sllaftl-ze  a new injection liner.  This applies only
                                   if the Injection liner is an integral part of the
                                  system.

                         10.3.6.3  Flush the GC column with solvent according to manu-
                                  facturer's instructions.

                         10.3.6.4  Break off a short portion (about 1  meter) of the
                                  column from the end near the injector; or replace GC
                                  column.  This action will cause a slight change in
                                  retention times.  Analyst may need  to redefine
                                  retention windows.

                         10.3.6.5  Prepare fresh CAL solutions, and repeat the  initial
                                 calibration step.

                         10.3.6.6 Clean the  MS ion source and rods (if a quadrupole).

                                           524.2-21
                                                            Rev.  4.1, August  1995

-------
                 10.3.6.7 Replace any components that allow analytes to come
                          into contact with hot metal surfaces.

                 10.3.6.8 Replace the MS electron multiplier, or any other
                          faulty components.

                 10.3.6.9 Replace the trap, especially when only a few com-
                          pounds fail the criteria in Sect. 10.3.5 while the
                          majority are determined successfully.  Also check
                          for gas leaks in the purge and trap unit as well as
                          the rest of the analytical system.

     10.4 Optional calibration for vinyl chloride using a certified gaseous
          mixture of vinyl chloride in nitrogen can be accomplished by the
          following steps.

          10.4.1 Fill the purging device with 25.0 ml (or 5-mL) of reagent
                 water or aqueous calibration standard.

          10.4.2 Start to purge the aqueous mixture.  Inject a known volume
                 (between 100 and 2000 fil) of the calibration gas (at room
                 temperature) directly into the purging, device with a gas
                 tight syringe.  Slowly inject the gaseous sample through a
                 septum seal at the top of the purging device at 2000 ^L/min.
                 If the injection of the standard is made through the aqueous
                 sample inlet port, flush the dead volume with several ml of
                 room air or carrier gas.  Inject the gaseous standard before
                 5 min of the 11-min purge time have elapsed.

          10.4.3 Determine'the aqueous equivalent concentration of vinyl
                 chloride standard, 1n /wj/L, injected with one of the
                 following equations:
                          5 ml samples, S
                         25 ml samples, S
                 
-------
ERRATA:  November 27, 1995
There is an error in the revision number.  This sheet corrects  that  error
Remove pages 525.2-1 and 525.2-2; and replace them with this  shee;:.
      METHOD 525.2      DETERMINATION OF ORGANIC COMPOUNDS IN DRINKING WATER
                        BY LIQUID-SOLID-EXTRACTION AND CAPILLARY COLUMN GAS
                        CHROMATOGRAPHY/MASS SPECTROMETRY
                                 Revision 2.0
                                        IF , •

                                        •sit'. .
            J.W.  Elchelberger, T.D. Behymer, H.L. Budde - Method 525.
            Revision 1.0,  2.0, 2.1 (1988)


            J.H.  Elchelberger, T.D; Behymer, and H.L.  Budde -. Method 525 1
            Revision 2.2 (July 1991)


            J.W.  Elchelberger, J.W. Munch^  and J.A.  Shoemaker
            Method  525.2  Revision 1.0 (February, 1994)

            J.W.  Munch  - Method 525.2, Revision 2.0  (1995)
                    NATIONAL EXPOSURE RESEARCH LABORATORY
                     OFFICE OF RESEARCH*AND  DEVELOPMENT
                    U.S. ENVIRONMENTAL  PROTECTION AGENCY
                           CINCINNATI,  OHIO  45268
                                          ,
                                          (•*%!_*•

                                   525.2-1 ~7

-------
                                 METHOD 525.2

             DETERMINATION OF ORGANIC COMPOUNDS  IN DRINKING WATER
               BY LIQUID-SOLID EXTRACTION AND CAPILLARY COLUMN  '
                     GAS CHROMATOGRAPHY/MASS SPECTROMETRY
1.     SCOPE AND APPLICATION

      1.1   This is a general purpose method that provides procedures  for
            determination of organic compounds  in finished drinking water,
            sourbe water, or drinking water in  any treatment stage.  The
            method is applicable to a wide range of organic compounds  that  are
            efficiently partitioned from the water sample onto a C18 organic
            phase chemically bonded to a solid  matrix  in a disk or cartridge,
            and sufficiently volatile and thermally stable for gas chromatog-
            raphy.  Single-laboratory accuracy  and precision data have been
            determined with two instrument systems using both disks and car-
            tridges for most of the following compounds:
           Analvte

          Acenaphthylene
          Alachlor
          Aldrin
          Ametryn
          Anthracene
          Atraton
          Atrazine
          Benz[a]anthracene
          Benzo[b;
          Benzo[k;
          BenzoJX
f 1 uoranthene
fluoranthene
pyrene
          Benzo[g,h,i]perylene
          Bromacil
          Butachlor
          Butyl ate
          Butylbenzylphthalate
          Carboxin
          Chlordane components
            Al pha-chlordane
            Gamma-chlordane
            Trans nonachlor
          Chlorneb
          Chiorobenzilate
          Chlorpropham
          Chlorothalonil
          Chlorpyrifos
          2-Chlorobiphenyl
152
269
362
227
178
211
215-
228
252
252
.252
276
260
311
217
312
235

406
406
440
206
324
213
264
349
188
                                 Chemical Abstracts Service
                                 	Registry Number
  208-
15972-
  309-
  834-
  120-
 1610-
 1912-
   56-
  205-
  207-
   50-
  191-
  314-
23184-
 2008-
   85-
 5234-

 5103-
 5103-
39765-
 2675-
  510-
  101-
 1897-
 2921-
 2051-
96-8
60-8
00-2
12-8
12-7
17-9
24-9
55-3
82-3
08-9
32-8
24-2
40-9
66-9
41-5
68-7
68-4

71-9
74-2
80-5
77-6
15-6
21-3
45-6
88-2
60-7
                                    525.2-2

-------
ERRATA - Nov. 27, 1995
There is an error in the analyte list.   This sheet corrects that error.   Remove pages
551.1-3 and 551.1.-4 and replace them with this sheet.

                                     Hexachlorocyclopentadiene       77.47-4
                                     Lindane (gamma-BHC)             58-89-9
                                   '-Jejtolachlor .   :;,s     ;       51218-45-2
                                     Metribuzin     ""~"           21087-64-9
                                     Methoxychlor                    72-43-5
                                     Simazine                       122-34-9
                                     Trifluralin                   1582-09-8

             1.2  This  analyte  list  includes twelve commonly  observed chlorination
                 disinfection  byproducts  (3,4),  eight  commonly  used  chlorinated
                 organic  solvents and  sixteen  halogenated  pesticides and  herbicides.

             1.3  This  method is  intended  as a  stand-alone  procedure  for either the
                 analysis  of only the  trihalomethanes  (THMs) or for  all the
                 chlorination  disinfection,  by-products  (DBFs) with the chlorinated
                 organic  solvents or as a procedure for the  total analyte  list  The
                 dechlorination/preseryation technique presented  in  section 8  details
                 two different dechlorinating  agents.  Results  for the THMs and the
                 eight  solvents may be obtained  from the analysis of samples
                 employing either dechlorinating agent. (Sect.  8.1.2)

            1.4  After  an  analyte has been  identified "and  quantitated in an unknown
                 sample with the primary GC column  (Sect.  6.9.2.1) qualitative
                 confirmation of results is strongly recommended by  gas
                 chromatography/mass spectrometry  (GC/MS)  (10,11), or by GC analysis
                 using a dissimilar column  (Sect. 6.9.2.2).          ;

            1.5  The experimentally determined method detection limits (MDLs)  (121
               .  for the above listed analytes are  provided in Tables 2 and 8
                 Actual MDL values will vary according to the particular matrix
                 analyzed and the specific instrumentation employed.

            1.6  This method is restricted to use by or under the supervision of
                 analysts experienced in the use of GC .and in the interpretation  of
                 gas chromatograms.   Each analyst must demonstrate the ability to
                 generate acceptable results with.this method using the procedure
                 described in Sect.  at1*.

            1.7  Methyl-t-butyl ether (MTBE) is recommended as  the primary extraction
                 solvent in this method since it effectively extracts all  of  the
                 target analytes listed In Sect.  1.1.  However,  due to safety
                 concerns associated with  MTBE  and the current  use of pentane  by  some
                 laboratories  for certain  method analytes,  pentane is offered  as  an
                 optional  extraction solvent for all analytes except  chloral  hydrate.
                 If  project requirements  specify the analysis of chloral hydrate,
                 MTBE must be used as  the  extracting solvent.  This method  includes
                 sections  specific for  pentane  as an optional solvent.

      2.   SUMMARY  OF  METHOD

           2.1   A 50 ml sample aliquot is extracted with 3 mL of MTBE or 5 ml  of
                 pentane.   Two jA. of the extract  1s  then injected  into a GC equipped
                                          551.1-3


                                                         "-     Rev. 1.0, August 1995

-------
          with a fused silica capillary column and linearized electron capture
          detector for separation and analysis.  Procedural  standard
          calibration is used to quantitate method analytes.

     2.2  A typical sample can be extracted and analyzed by this method in 50
          min for the chlorination by-products/chlorinated solvents and 2 hrs.
          for the total analyte list.  Confirmation of the eluted compounds
          may be obtained using a dissimilar column (6.9.2.2) or by the use of
          GC-MS.  Simultaneous confirmation can be performed using dual
          primary/confirmation columns installed in a single injection port
          (Sect. 6.9.3) or a separate confirmation analysis.

3.   DEFINITIONS

     3.1  INTERNAL STANDARD (IS)'— A pure analyte(s) added to a sample,
          extract, or standard solution in known amount(s) and used to measure
          the relative responses of other method analytes and surrogates that
          are components of the same sample or solution.  The internal
          standard must be an analyte that is not a sample component.

     3.2  SURROGATE ANALYTE (SA) — A pure analyte(s), which is extremely
          unlikely to be found in any sample, and which is added directly to  a
          sample aliquot in known amount(s) before extraction or other
          processing and is measured with the same procedures used to measure
          other sample components. The purpose of a surrogate analyte is to
          monitor method performance with each sample.

     3.3  LABORATORY DUPLICATES (LD1 and LD2) -- Two sample aliquots, taken  in
          the laboratory from a single sample bottle, and analyzed separately
          with  identical procedures.  Analyses of LD1 and LD2 indicate
          precision associated with laboratory procedures, but not with sample
          collection, preservation, or storage procedures.   This method
          cannot utilize laboratory duplicates since sample extraction must
          occur in the sample vial and sample transfer  is not possible due to
          analyte volatility.

     3.4  FIELD DUPLICATES (FD1 and FD2) -.-.Two separate samples collected at
          the same time and plate under identical circumstances and treated
          exactly the same throughout field and laboratory procedures.
          Analyses of FD1 and FD2 give a measure of the precision associated
          with  sample collection, preservation and storage, as well as with
          laboratory procedures.  Since laboratory duplicates cannot be
          analyzed, the collection and analysis of field duplicates for this
          method  is critical.

     3.5  LABORATORY REAGENT BLANK  (LRB) — An aliquot  of reagent water, or
          other blank matrix, that  1s treated  exactly as a  sample including
          exposure to all glassware, equipment, solvents, reagents,  internal
          standards, and surrogates that are used with  other samples.  The LRB
          is used to determine  if method analytes or other  interferences are
          present  in the laboratory environment,  the reagents, or the
          apparatus.
                                    551.1-4

-------
ERRATA - Nov.  27,  iyy5           .--,-•-y. „>-•
There are several  errors in Section 7.1.7.1 and 7.1.7.4.  This sheet corrects those errors
Remove pages 551.1-11 and 551.1-12 and replace them with this sheet.


                          temperature to 400°C and hold for 30 min.--Store in a capped
                          glass  bottle not in a plastic container.

                   7.1.7  Sample  Preservation Reagents

                          7.1.7.1   Phosphate buffer - Used  to  lower the  sample  matrix
                                    pH to 4.8-5.5  in order to  inhibit base  catalyzed
                                    degradation of the haloacetonitriles  (7),  some  of
                                    the chlorinated solvents,  and  to standardize the pH
                                    of all samples.  Prepare a  dry homogeneous mixture
                                    of 1% Sodium Phosphate,  dibasic (Na2HP04)/99%
                                    Potassium Phosphate, monobasic (KH2P04)  by weight
                                    (example:  2 g Na2HP04 and  198 g KH2P04  to yield a
                                    total weight of 200 g)   Both of these buffer salts
                                    should be in granular form  and of ACS grade  or
                                    better.  Powder would be ideal  but would require
                                    extended cleanup time as outlined below in Sect.
                                    7.1.7.5 to allow for buffer/solvent settling.

                          7.1.7.2  Arnmprfium Chloride, NH4C1, ACS  Reagent Grade.  Used
                                    to convert free chlorine to monochloramine.
                                   Although this  is not the traditional dechlorination
                                   mechanism, ammonium chloride is  categorized  as  a
                                   dechlprinating" agent in this method.

                          7.1.7.3  Sodfum Sulfite, Na2S03, ACS Reagent Grade.   Used as
                                   a dechlorinating agent for chloral hydrate sample
                                    analysis.

                          7.1.7.4  To ||mplify the addition of 6  mg  of the
                                   dechlorinating agent to the 60 ml vial, the
                       • ...  .       dechlorinating salt is combined with the phosphate
                                   buffer as a homogeneous mixture.  Add 1.2 g  of  the
                                   appropriate dechlorinating agent  to 200 g of the
                                   phosphate buffer.  When 1 g of the buffer/
                                   decWForinating agent mixture are  added to the 60-mL
                                   sairiple vial, 6 mg of the dechlorinating agent are
                                   included reflecting an actual concentration of  100
                                   mg/L.  Two separate mixtures are prepared, one
                                   containing ammonium chloride and the other with
                                   sodium sulfite.
                          7.1.7.5  If background contaminants are detected in the salts
                                   listed in Sections 7.1.7.1 through 7.1.7.3, a
                                   solvent rinse cleanup procedure may be required.
                                   Ttifse contaminants may coelute with some of the high
                                   mpl,ecular weight herbicides and pesticides.  These
                                   salts cannot be muffled since they decompose when
                                   heated to 400°C.   This solvent rinsing procedure is
                                   applied to the homogeneous mixture prepared in Sect.
                                   7*1.7.4.
                                  •HI
                                  |i|    "" 551.1-11

                                  'i-iSlv
                                  -\± •$.*'
                                  0f-                          Rev.  1.0,  August  1995

-------

-------
                                   FOREWORD

     Environmental measurements are required to determine the quality of
ambient waters and the character of waste effluents.  The National Exposure
Research Laboratory - Cincinnati (NERL-Cincinnati) conducts research to:
     o  Develop and evaluate analytical methods to identify and measure the
        concentration of chemical pollutants in drinking waters, surface
        waters, groundwaters, wastewaters, sediments, sludges, and solid
        wastes.

     o  Investigate methods _for the identification and measurement of viruses,
        bacteria and other microbiological organisms in aqueous samples and to
        determine the responses of aquatic organisms to water quality.

     o  Develop and operate a quality assurance program to support the
        achievement of data quality objectives in measurements of pollutants
        in drinking water, surface water, groundwater, wastewater, sediment
        and solid waste.                •

     o  Develop methods and models to detect and quantify response in aquatic
        and terrestrial organisms exposed to environmental stressors and to
        correlate the exposure with effects on chemical and biological
        indicators.       ;i|;

     This publication, "Determination of Organic Compounds in Drinking Water
Supplement III," was prepared to gather together under a single cover a set of
15 new or improved laboratory analytical methods for organics compounds in
drinking water.  NERL-Cincinnati is pleased to provide this manual and believe
that it will be of considerable value to many public and private laboratories
that wish to determine organic compounds in drinking water for regulatory or
other reasons.           *;•-
                                        Alfred P. Dufour, Acting Director
                                        Aquatic Research Division
                                        National Exposure Research
                                        Laboratory - Cincinnati
                                      m

-------
                                   ABSTRACT
     Fifteen analytical methods for organic compounds in drinking water are
documented in detail.  Most of these methods were published as prior versions
in other methods manuals in this series.  The versions in Supplement III
provide corrections, minor technical enhancements, and editorial improvements
to the previously published analytical methods.  Several previously distrib-
uted but not formally published methods are also included.  Fourteen of the
fifteen methods utilize high resolution gas chromatography (GC) for separation
of analytes from each other and from other substances in the water sample.
One method employs high performance reverse phase liquid chromatography for
the separation.  Two methods utilize a mass spectrometer for the unambiguous
identification and measurement of the compounds separated by high resolution
GC.  These two methods are extremely versatile and have been single-laboratory
validated for a total of 194 individual compounds and 8 commercial product
mixtures.  Most methods have also been multi-laboratory validated although not
all possible analytes have been included in these studies.  Essentially all of
the major chlorine disinfection by-products that have been identified in
drinking water are included in the methods in this manual.
                                      iv

-------
                               TABLE  OF  CONTENTS

Method                 i
Number   Title                                             Revision     Page

         Foreward .  •  /	........ 1 ..      ill

         Abstract .  .  •	       iv

         Acknowledgment	 .....       vii

         Analyte - Method Cross Reference 	 .' 	      viii

         Introduction  ...	•......'.	        1

502.2    Volatile Organic Compounds in Water by Purge         2.1
         and Trap Capillary Column Gas Chromatography         !
                      *<
         with Photoionfzation and Electrolytic
         Conductivity Detectors in Series
                       I1';
504.1    1,2-Dibromoethfne (EDB), l,2-Dibromo-3-Chloro-       1.1
         propane (DBCP)^ and 1,2,3-Trichloropropane           i
         (123TCP) in Wa%r by Microextraction and Gas         !
         Chromatography |;                                    '  •

505      Analysis of Orga-hohal ide Pesticides and              2.1
         Commercial  Polychlorinated Biphenyl (PCB)
         Products in Water by Microextraction and Gas
         Chromatography   u

506      determination of Phthalate and Adi pate Esters        1.1
         in Drinking Water by Liquid-Liquid Extraction
         or Liquid-Solid Extraction and Gas
         Chromatography with Photoionization Detection

507      Determination of Nitrogen- and Phosphorus-           2.1
         Containing Pesticides in Water by Gas
         Chromatography with a Nitrogen-Phosphorus Detector

508      Determination of Chlorinated Pesticides in Water     3.1
         by Gas Chromatography with an Electron Capture
         Detector

508.1    Determination of Chlorinated Pesticides,             2.0
         Herbicides, and Organohalides by Liquid-Solid
         Extraction and Electron Capture Gas Chromatography

509      Determination of Ethylene Thiburea (ETU) in          1.1
         Water using Gas Chromatography with a
         Nitrogen-Phosphorus Detector

-------
                         TABLE OF CONTENTS (Continued)
 Method
 Number

 515.1


 515.2



 524.2



 525.2



 531.1



 551.1
552.2
 Title
 Determination of Chlorinated Acids' in Water by Gas
 Chromatography with an Electron Capture Detector

 Determination ,of Chlorinated Acids in Water
 using Liquid-Solid Extraction and Gas
 Chromatography with an Electron Capture Detector

 Measurement of Purgeable Organic Compounds  in
 Water by Capillary Column Gas Chromatography/Mass
 Spectrometry

 Determination of Organic Compounds in Drinking
 Water by Liquid-Solid  Extraction and  Capillary
 Column  Gas  Chromatography/Mass  Spectrometry

 Measurement of N-Methylcarbamoyloximes  and
 N-Methylcarbamates  in  Water  by  Direct Aqueous
 Injection HPLC with  Post Column  Derivatization

 Determination  of Chlorination Disinfection
 Byproducts,  Chlorinated  Solvents,  and Halogenated
 Pesticides/Herbicides  in  Drinking  Water by
 Liquid-Liquid  Extraction  and  Gas Chromatography
with  Electron-Capture  Detection
Revision

   4.1
                                                                Page
                                                       in
Determination of HaToacetic Ac:ds and Dalapon
Drinking Water by Liquid-Liquid Extraction,
Derivatization and Gas Chromatography with Electron
Capture Detection
   1.1
   4.1
   2.0
  3.1
                                                              1.0
  1.0
                                     VI

-------
                                ACKNOWLEDGMENT
      The many, past and present researchers and authors who contributed to the
current status of the methods in supplement III are recognized oh the title
pages of the 15 methods in Supplement III. ! Jean W. Munch deserves special
recognition for Supplement III because she critically read every method and
incorporated numerous corrections, technical  enhancements, and editorial
improvements into each of them.  Thomas 0. Behymer acquired the precision and
accuracy data for the Aroclors in Method 525.2, proof read the next-to-last
drafts of all the methods, and provided numerous additional corrections.
Special thanks is due to Diane Schirmann who processed many of the final
corrections, placed all 15 methods and the introductory material in the
standard format, and prepared the camera-ready copies for printing.  The
authors would like to thank the many others who reviewed previous and current
versions of these methods and provided comments, suggestions and corrections.
                                      VII

-------
                        ANALYTE - METHOD CROSS REFERENCE
    ANALYTE

 Acenaphthylene
 Acetone
 Acifluorfen
 Acrylonitrile
 Alachlor
 Aldicarb
 Aldicarb sulfone
 Aldicarb sulfoxide
 Aldrin
 Ally! chloride
 Ametryn
 Anthracene
 Atraton
 Atrazine
 Baygon
 Bentazon
 Benzene
 Benz[a]anthracene
 Benzo[b]fluoranthene
 Benzo[k]fluoranthene
 Benzo[a]pyrene
 Benzo[g,h,i]perylene
 Bis (2-ethylhexyl)  phthalate
 Bis (2-ethylhexyl)  adipate
 Bromobenzene
 Bromacil
 Bromochloroacetic  acid
 Bromochloroacetoni tri1e
 Bromochloromethane
 Bromodichloroacetic acid
 Bromodi chloromethane
 Bromoform
 Bromomethane
 Butachlor
 2-Butanone
 Butyl ate
 Butyl benzylphthalate
 n-Butylbenzene
 sec-Butyl benzene
 tert-Butylbenzene
 Carbaryl
 Carboxin
 Carbofuran
 Carbon disulfide
 Carbon tetrachloride
 Chloramben
 Chloral Hydrate
Chlordane
                    METHOD  NO.
                 515.1,

505, 507, 508.1, 525.2,



       505, 508, 508.1,

                   507,

                   507,
505, 507, 508.1, 525.2,

                 515.1,
                 502.2,
                 502.2,
            507,  525.2,
          502.2,  524.2,
                  507,
                  506,
                502.2,
                502.2,
                502.2,

                  507,
         502.2, 524.2,
 525
 524
 515
 524
 551
 531
 531
 531
 525
 524
 525
 525
 525
 551
 531.1
 515.2
 524
 525
 525
 525
 525
 525
  506
  506
                 502.2,
          502.2,  524.2,
                 502.2,
            507,  508.1, 525.2
524
551
552,
551.
551,
552.
524.
551.
524.
524.
525.
525.
524:2
524,
524.
531.
525.
531.
524.
551.
                       515.1
                       551.1
                    505, 508
                                     vm

-------
    ANALYTE
                                                                   METHOD NO.
   Alpha-chlordane
   Gamma-chlordane
   Trans nonachlor
 Chloroacetqnitrile
 Chloroberizene
 Chiorobenzi late
 2-Chlorobiphenyl
 Chlorodibromoacetic acid
 1-Chlorobutane
 Chloroethane
 Chloroform
 Chloromethane
 Chloroneb
 Chloropicrin
 Chlorothalonil
 2-Chlorotoluene
 4-Chlorotoluene
 Chlorpropham
 Chlorpyrifos
 Chrysene
 Cyanazine
 Cycloate
 Dacthal(DCPA)
 2,4-D
 Dalapon
 2,4-DB
 DCPA acid  metabolites
 4,4'-DDD
 4,4'-DDE
 4,4'-DDT
 Diazinon
 Dibenz[a,h]anthracene
 Dibromoacetic  acid
 Dibromoacetonitri1e
 Dibromochloromethane
 1,2-Di bromo-3-chloropropane
 1,2-Dibromoethane
 Dibromomethane
 Dicamba
 Dichloroacetic  acid
 DiChloroacetonitrile
 1,2-Dichlorobenzene
 1,3-Dichlorobenzene
 1,4-Dichlorobenzene
 3,5-Dichlorobenzoic acid
 trans-l,4-Dichloro-2-butene
 1,1-Dichloroethane
 1,2-Dichloroethane
 1,1-Dichloroethene
cis-1,2-Dichloroethene
505, 508,
505, 508,



508,




502.2,

508,

508,





508.1,



515.1,

508,
508,
508,
508,




502.2,
502.2, 504.1,
502.2, 504.1,













508.1, 525.2
508.1, 525.2
525.2
524.2
502.2, 524.2
508.1, 525.2
525.2
552.2
524.2
502.2, 524.2
524.2, 551.1
502.2, 524.2
508.1, 525.2
551.1
508.1, 525.2
502.2, 524.2
502.2, 524.2
507, 525.2
525.2
525.2
525.2, 551.1
507, 525.2
525.2
515.1, 515.2
515.2, 552.2
515.1, 515.2
508.1, 515.1
508.1, 525.2
508.1, 525.2
508.1, 525.2
507, 525.2
525.2
552.2
551.1
524.2, 551.1
524.2, 551.1
524.2, 551.1
502.2, 524.2
515.1, 515.2
552.2
551.1
502.2, 524.2
502.2, 524.2
502.2, 524.2
515.1, 515.2
524.2
502.2, 524.2
502.2, 524.2
502,2, 524.2
502.2, 524.2
                                      IX

-------
   ANALYTE
                   METHOD NO.
trans-1,2-Di chloroethene
Di chlorodi f1uoromethane
1,2-Di chloropropane
1,3-Di chloropropane
2,2-Dichloropropane
1,1-Di chloropropene
1,1-Di chloropropanone
ci s-1,3-Di chloropropene
trans-1,3-Di chloropropene
1,1-Di chloro-2-propanone
Dichloroprop
Di-n-butyl phthalate
Di-n-octyl phthalate
2,3-Dichlorobiphenyl
Dichlorvos
Dieldrin
Diethyl ether
Diethyl phthalate
Di(2-ethylhexyl)adi pate
Di(2-ethylhexyl)phthalate
Dimethyl phthalate
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Dinoseb
Diphenamid
Disulfoton
Disulfoton sulfone
Disulfoton sulfoxide
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Endrin ketone
EPTC
Ethoprop
Ethyl benzene
Ethyl methacrylate
Ethylene thiourea
Etridiazole
Fenamiphos
Fenarimol
Fluorene
Fluridone
Heptachlor
Heptachlor Epoxide
2,2',3,3',4,4',6-Heptachloro-
  biphenyl
Hexachlorobenzene
Hexachlorobutadiene
                 502.2,  524.2
                 502.2,  524.2
                 502.2,  524.2
                 502.2,  524.2
                 502.2,  524.2
                 502.2,  524.2
                        524.2
                 502.2,524.2
                 502.2,  524.2
                        551.1
                 515.1,  515.2
                   506,  525.2
                          506
                        525! 2
                   507,  525.2
       505,  508,  508.1,  525.2
                        524.2
                   506,  525.2
                        525.2
                        525.2
                   506,  525.2
                        525.2
                        525.2
                 515.1,  515.2
                   507,  525.2
                   507,  525.2
                   507,  525.2
                   507,  525.2
            508,  508.1,  525.2
            508,  508.1,  525.2
            508,  508.1,  525.2
505, 508, 508.1,  525.2,  551.1
     508, 508.1,  525.2,  551.1
                        551.1
                   507,  525.2
                   507,  525.2
                 502.2,  524,2
                        524.2
                          509
            508,  508.1,  525.2
                   507,  525.2
                   507,  525.2
                        525.2
                   507,  525.2
505, 508, 508.1,  525.2,  551.1
505, 508, 508.1,  525.2,  551.1

                        525.2
505, 508, 508.1,  525.2,  551.1
                 502.2,  524.2

-------
   ANALYTE
       METHOD NO.
Hexachlorocyclopentadi ene
2,2',4,4',5,6'-Hexachloro-
  biphenyl
Hexachlorocyclohexane, alpha
Hexachlorocyclohexane, beta
Hexachlorocyclohexane, delta
Hexachlorocyclopentadiene
Hexachloroethane
2-Hexanone
Hexazinone
HCH-alpha
HCH-beta
HCH-delta
HCH-gamma (lindane)
3-Hydroxycarbofuran
5-Hydroxydicamba
Indeno[l,2,3,c,d]pyrene
Isophorone
Isopropylbenzene
4-Isopropyltoluene
Lindane  (gamma-BHC)
Merphos
Methacrylonitrile
Methiocarb
Methomyl
Methoxychlor
Methylacrylate
Methylene chloride
Methyl iodide
Methylmethacrylate
Methyl paraoxon
4-Methyl-2-pentanone
Methyl-t-butyl-ether
Metolachlor
Metribuzin
Mevinphos
MGK 264
Molinate
Monobromoacetic acid
Monochloroacetic acid
Naphthalene
Napropamide
Nitrobenzene
4-Nitrophenol
2-Nitropropane
cis-Nonachlor
Norflurazon
2,2',3,3',4,5',6,6'-Octa-
  chlorobiphenyl
505,  508.1,  551.1

            525.2






507,
508,
508,
508,
508,

515.1,


502.2,

505, 525.2,
507,



505, 508, 508.1, 525.2,

502.2,


507,


507, 508.1, 525.2,
507, 508.1, 525.2,
507,
507,
507,


502.2,
507,




507,
525.2
525.2
525.2
525.2
524.2
524.2
525.2
508.1
508.1
508.1
508.1
531.1
515.2
525.2
525.2
524.2
524.2
551.1
525.2
524.2
531.1
531.1
551.1
524. -2
524.2
524.2
524.2
525.2
524.2
524.2
551.1
551.1
525.2
525.2
525.2
552.2
552.2
524.2
525.2
524.2
515.1
524.2
505
525.2
            525.2
                                      XI

-------
    ANALYTE
                                                                    METHOD NO.
 Oxamyl
 Pebulate
 2,2',3',4,6-Pentachloro-
   biphenyl
 Pentachloroethane
 Pentachlorophenol
 cis-Permethrin
 Trans-Permethrin
 Phenanthrene
 Picloram
 Prometon
 Prometryn
 Pronamide
 Propachlor
 Propazine
 Propionitrile
 Propylbenzene
 n-Propylbenzene
 Pyrene
 Simazine
 Simetryn
 Stirofos
 Styrene
 2,4,5-T
 2,4,5-TP
 Tebuthiuron
 Terbacil
 Terbufos
 Terbutryn
 2,2',4,4'-Tetrachlorobiphenyl
 1,1,1,2-Tetrachloroethane
 1,1,2,2-Tetrachloroethane
 Tetrachloroethene
 Tetrachloroethylene
 Tetrahydrofuran
 Toluene
 Toxaphene
 Triademefon
 Tribromoacetic  acid
 Trichloroacetic  acid
 Trichloroacetonitrile
 1,2,3-Tri chlorobenzene
 1,2,4-Trichlorobenzene
 1,1,1-Trichloroethane
 1,1,2-Tri chloroethane
Trichloroethene
Trichloroethylene
Tri chlorof1uoromethane
 1,1,1-Tri chloro-2-propanone
 1,2,3-Tri chloropropane
                 531.1
            507, 525.2
   515.1, 515.2,
     508, 508.1,
     508, 508.1,

          515.1,
            507,
            507,
            507,
     508, 508.1,
            507,
505, 507, 508.1,
            507,
            507,
          502.2,
          515.1,
          515.1,
            507,
            507,
            507,
            507,

          502.2,
          502.2,
          502.2,
          502.2,
505,  508,  508.1,
            507,
          505.2,
          505.2,
   502.2,  524.2,
   502.2,  524.2,
          505.2,
          502.2,

   504.1,  524.2,  551.1
 525.2
 524.2
 525
 525
 525
 525
 515
 525
 525
 525
 525
 525
 524
 502
 524
 525
 525
 525
 525
 524
 515
 515.2
 525.2
 525.2
 525
 525
 525
 524
 524
 524
 551
 524
 524
 525
 525
 552
 552
 551
 524
524
551
551
524
551
524
551
                                      xn

-------
    ANALYTE                                                          METHOD NO.

 2,4,5-Trichlorobiphenyl                                          502.2, 525.2
 Tricyclazole                                                       507, 525.2
 Trifluralin                                          508, 508.1, 525.2, 551.1
 1,2,4-Trimethylbenzene                                          502.2, 524.2
 1,3,5-Trimethylbenzene                                          502.2, 524.2
 Vernolate                                                          507, 525.2
 Vinyl  chloride                                                   502.2, 524.2
 0-Xylene                                                         505.2, 524.2
 m-Xylene                                                      •   502.2, 524.2
 p-Xylene                                                         502.2, 524.2
 Aroclor 1016                                           505, 508, 508.1, 525.2
 Aroclor 1221                                           505, 508, 508.1, 525.2
 Aroclor 1232                                           505, 508, 508.1, 525.2
 Aroclor 1242                                           505, 508, 508.1, 525.2
•Aroclor 1248                                           505, 508, ,508.1, 525.2
 Aroclor 1254                                           505, 508, 508.1, 525.2
 Aroclor 1260                                           505, 508, 508.1, 525.2
                                     xm

-------

-------
                                  INTRODUCTION


                      William L.  Budde and Jean W. Munch
     The purpose of this third supplement to "Methods for the Determination of
Organic Compounds  in Drinking Water" is to provide corrections, minor tech-
nical enhancements, and editorial improvements to some previously published
analytical methods and to document several significantly enhanced or pre-
viously unpublished methods.  Some of these modifications were described in
"Technical Notes on Drinking Water Methods", EPA/600/R-94/173, October, 1994
and these method changes have been incorporated into the body of the methods
in this supplement.  All methods in this supplement are written'in a format
specified by the United States Environmental Protection Agency's Environmental
Monitoring Management Council.

     As in other manuals in this series,, each of the methods in Supplement III
was intended to stand alone, that is, each method may be removed from the
manual, photocopied, inserted into another binder, and used without loss of
information.  The  stand-alone character of the methods comes at some cost of
duplication of material, but the authors believe that the added bulk of the
methods is a small price to pay for the flexibility of the format.

     All the methods in supplement III have been given new dates and version
numbers or slightly modified method numbers to distinguish them from pre-
viously published  versions. A change in the revision number indicates a
relatively small modification to the method and a change in the method number
usually indicates  a relatively larger change in the method.  The cover page of
each method gives  the title, method number, revision, and date, and also lists
the  previous versions, previous authors, and dates if they are known.  The
purpose of this very brief method history is to assist users who may have
older versions in  their files in understanding the chronological relationship
of methods as technical improvements were made over the years.  It also gives
due credit to previous authors who  contributed to the development of the
methods in this and previous manuals in this series.  Unless otherwise
indicated, all authors were direct Federal employees of the U. S.
Environmental Protection Agency at the time of their contributions.

     Some methods  in supplement III utilize the liquid-solid extraction
technology which was pioneered by the USEPA in the original Method 525 in
1988.  The scientifically correct term liquid-solid extraction (LSE), in which
both phases of the equilibrium partition process are named, is used throughout
the manual in place of the misleading commercial term "solid phase
extraction".  Colloquial lab terms such as "clean-up" and "spike" are replaced
by "sample preparation" or "interference separation" and "fortified"
respectively.

     While the title of supplement III, and previous manuals in  this series,
specifies drinking water,  these methods will  very likely be applicable to
other aqueous matrices including surface water,  ground water,  beverages, and

                                       1

-------
waste water.  However since some methods have been tested with only reagent
water and/or drinking water, caution is needed when applying these methods to
matrices other than reagent or drinking water.  One exception is Method 524.2
which has been tested in a large multi-laboratory validation study with a
variety of aqueous matrices.

     During 1992 the USEPA and the American Society for Testing and Materials
(ASTM) Committee D-19 on Water jointly conducted a multi-laboratory study of
an ASTM version of Method 524.2, revision 3.0 using 68 of the volatile organic
compound analytes. Over 40 volunteer laboratories participated in the study to
characterize the performance of Method 524.2 in terms of accuracy, precision,
and detection limits.  Analyses were conducted using fortified reagent water,
drinking water, ground water, several industrial waste waters, and a simulated
hazardous waste site aqueous leachate. Fortified analyte concentrations ranged
from 0.2 /ig/L to 80 /jg/L and generally excellent accuracy and precision was
reported.  Full details of that study will be published by the ASTM.

     The methods in supplement III are listed below along with a comment for
each that gives the previous version of the method, the citation and date of
publication of the previous version, and the highlights of the changes in the
version in supplement III.
Method in SUDP. Ill

502.2   rev.  2.1
                               Comments

                 Rev. 2.0 was published in EPA/600/4-88/039 in Dec.
                 1988 and July 1991.  Rev. 2.1 is modified to specify
                 conditions under which alternative trapping materials
                 may be used and instructions are clarified for sample
                 preservation and dechlorination.  Conditions which do
                 not require a photoic.,ization detector are specified.
504.1   rev.  1.1
505
rev.   2.1
506
rev.  1.1
Method 504.1 improves Method 504 rev. 2.0 which was
published in EPA/600/4-88/039 in Dec. 1988 and July
1991.  In 504.1 changes are made to the sampling
procedures, the holding time, and the compound 1,2,3-
trichloropropane is added to the analyte list.
Cautions are included on the frequent coelution of
ethylene dibromide and bromodichloromethane.

Rev. 2.0 was published in EPA/600/4-88/039 in Dec.
1988 and July 1991.  Rev. 2.1 is modified to  remove
alternative detectors except mass spectrometry for
qualitative confirmation and to provide additional
instructions on the measurement of multi-component
mixtures.

Rev. 1.0 was published in EPA/600/4-90/020 (Supp. I)
in July, 1990.   Rev. 1.1 is modified to correct
errors in the method summary.

-------
507     rev.  2.1
508     rev.  3.1
508.1   rev.  2.0
509     rev. 1.1
515.1   rev. 4.1
515.2   rev.  1.1
524.2   rev. 4.1
525.2   rev.  2.0
 Rev.  2.0  was  published  in  EPA/600/4-88/039  in Dec.
 1988  and  July 1991.  Rev. 2.1  is  modified  to  remove
 mercuric  chloride  as a  preservative.   Data  tab\es are
 reorganized for  clarity and  addition  of method
 detection limits.  Alternative detectors  are
 eliminated except  mass  spectrometry for qualitative
 confirmation.

 Rev 3.0 was published in EPA/600/4-88/039 in Dec.
 1988  and  July 1991.   Rev.  3.1  is modified to remove
 mercuric  chloride  as a  preservative.   Data  tables are
 reorganized for  clarity and  addition  of method
 detection limits.  Alternative detectors  are
 eliminated except  mass  spectrometry for qualitative
 confirmation.

 Method 508.1  is  a  significant  improvement to Method
 508.   It  employs the liquid-solid extraction
 technology of Method 525.2 and has an analyte  list
 consisting of many Method  507  and Method  508
 substances including the former  commercial  Aroclor
 mixtures.

 Method 509 is a  single  analyte method for the
 pesticide metabolite ethylenethiourea. This method
 is derived from  national pesticide survey Method 6.

.Rev.  4.0  was  published  in  EPA/600/4-88/039 in  Dec.
 1988  and  July 1991.   Rev.  4.1  is modified to remove
 mercuric  chloride  as a  preservative.   Data tables  are
 reorganized  for  clarity and  addition  of method
 detection limits.  Trimethylsilyldiazomethane  (TMSD)
 is added  as  an alternative methylating agent.

 Rev.  1.0  was  published  in  EPA/600/R-92/129 (Supp.  II)
 in August, 1992.  Rev.  1.1 is modified to include
 trimethylsilyldiazomethane (TMSD) as  an  alternative
 methylating  agent.

 Rev.  4.0  was  published  in  EPA/600/R-92/129 (Supp.  II)
 in August, 1992.  Rev.  4.1 is modified to specify
 conditions under which  alternative trapping materials
 may be used  and  instructions are clarified for sample
 preservation  and dechlorination.  The quality
 assurance section  is clarified and data  for two
 analytes  are  added.

 Method 525.2, rev. 2.0 is  an improvement  to Method
 525.1, rev.  2.2  which was  published in EPA/600/4-
 88/039 in July,  1991.  Method 525.2 includes criteria
 for judging  the  equivalency of alternative liquid-
 solid extraction cartridges and  disks.  The elution

-------
531.1   rev. 3.1
551.1   rev.  1.0
552.2   rev. 1.0
solvent is modified and data are included for
additional analytes including all former commercial
Aroclor mixtures.

Rev. 3.0 was published in EPA/600/4-88/039 in Dec.
1988 and July 1991.  Rev. 3.1 is modified to remove
the requirement to freeze the samples.  Data tables
are revised for clarity and method detection limits
are included.

Method 551.1 is an improvement to Method 551 which
was published in EPA/600/4-90/020 (Supp. I) in July
1990.  Pentane is included as an alternative solvent
for some analytes, the analyte list is expanded, and
a new sample preservation technique is used.

Method 552.2 is similar to Method 552 which was
published in EPA/600/4-90/020 (Supp. I) in July,
1990.  Method 552.2 uses acidic methanol for
methylation instead of diazomethane and expands the
analyte list.

-------
METHOD 502.2
VOLATILE ORGANIC COMPOUNDS IN WATER BY PURGE AND TRAP
CAPILLARY COLUMN GAS CHROMATOGRAPHY WITH PHOTOIONIZATION
AND ELECTROLYTIC CONDUCTIVITY DETECTORS IN SERIES
                            Revision 2.1




                     Edited by J.W.  Munch (1995)




     R.W. Slater, Jr. and J.S. Ho - Method 502.2, Revision 1.0 (1986)

     J.S. Ho - Method 502.2, Revision 2.0 (1989)
                NATIONAL EXPOSURE RESEARCH LABORATORY
                 OFFICE OF RESEARCH AND DEVELOPMENT
                U.S. ENVIRONMENTAL PROTECTION AGENCY
                       CINCINNATI, OHIO 45268
                               502.2-1

-------
                                 METHOD 502.2

            VOLATILE ORGANIC COMPOUNDS IN WATER BY PURGE AND TRAP
          CAPILLARY  COLUMN  GAS  CHROMATOGRAPHY WITH  PHOTOIONIZATION
              AND ELECTROLYTIC CONDUCTIVITY DETECTORS IN SERIES
1.   SCOPE AND APPLICATION

     1.1  This is a general purpose method for the identification and
          simultaneous measurement of purgeable volatile organic compounds in
          finished drinking water, raw source water,  or drinking water in any
          treatment stage (1-3).  The method is applicable to a wide range of
          organic compounds, including the four trihalomethane disinfection
          by-products, that have sufficiently high volatility and low water
          solubility to be efficiently removed from water samples with purge
          and trap procedures.  The following compounds can be determined by
          this method.
                Analvte

          Benzene
          Bromobenzene
          Bromochloromethane
          Bromodi chloromethane
          Bromoform
          Bromomethane
          n-Butylbenzene
          sec-Butyl benzene
          tert-Butylbenzene
          Carbon tetrachloride
          Chlorobenzene
          Chloroethane
          Chloroform
          Chloromethane
          2-Chlorotoluene
          '4-Chlorotoluene
          Di bromochloromethane
          1,2-Di bromo-3-chloropropane
          1,2-Dibromoethane
          Dibromomethane
          1,2-Di chlorobenzene
          1,3-Di chlorobenzene
          1,4-Di chlorobenzene
          Dichlorodif1uoromethane
          1,1-Dichloroethane
          1,2-Dichloroethane
          1,1-Dichloroethene
          cis-1,2-Dichloroethene
          trans-1,2-Di chloroethene
          1,2-Di chloropropane
Chemical  Abstract Services
       Registry Number

          71-43-2
         108^86-1
          74-97-5
          75-27-4
          75-25-2
          74-83-9
         104-51-8
         135-98-8
          98-06-6
          56-23-5
         108-90-7
          75-00-3
          67-66-3
          74-87-3
          95-49-8
         106-43-4
         124-48-1
          96-12-8
         106-93-4
          74-95-3
          95-50-1
         541-73-1
         106-46-7
          75-71-8
          75-34-3
         107-06-2
          75-35-4
         156-59-4
         156-60-5
          78-87-5
                                   502.2-2

-------
           1,3-Dichlpropropane                            142-28-9
           2,2-Dichloropropane.                            590-20-7
           1,1-Dichloropropene                            563-58-6
           cis-l,3-Dichloropropene                      10061-01-5
           trans-1,3-Dichloropropene                    10061-02-6
           Ethylbenzene                                   100-41-4
           Hexachlorobutadiene                             87-68-3
           Isopropylbenzene-                                98-82-8
           4-Isopropyltoluene                              99-87-6
           Methylene  chloride                              75-09-2
           Naphthalene                                     91-20-3
           Propylbenzene                                  103-65-1
           Styrene                                        100-42-5
           1,1,1,2-Tetrachloroethane                      630-20-6
           1,1,2,2-Tetrachloroethane                       79-34-5
           Tetrachloroethene                              127-18-4
           Toluene ,                                       108-88-3
;           1,2,3-Trichlorobenzene                 .         87-61-6
           1,2,4-Trichlorobenzene                         120-82-1
           1,1,1-Trichloroethane                           71-55-6
           1,1,2-Trichloroethane                           79-00-5
           Trichloroethene                                 79-01-6
           Trichlorofluoromethane                          75-69-4
           1,2,3-Trichloropropane                          96-18-4
           1,2,4-Trimethyl benzene                          95-63-6
           1,3,5-Trimethylbenzene                         108-67-8
           Vinyl  chloride                                  75-01-4
           o-Xylene                                        95-47-6
           m-Xylene                                       108-38-3
           p-Xylene                                       106-42-3

      1.2   This  method is applicable to the determination  of total
           trihalomethanes, and other volatile organic compounds  (VOCs).  Method
           detection  limits (MDLs) (4)  are compound  and  instrument dependent
           and vary from approximately  0.01-3.0 fig/I.  The applicable
           concentration range of this  method is also compound and instrument
           dependent  and is approximately 0.02 to 200 //g/L.   Analytes  that  are
           inefficiently purged from water will  not  be detected  when present at
           low concentrations, but they can be measured  with acceptable
           accuracy and precision when  present in sufficient amounts.

      1.3   Two of the three isomeric xylenes may not be  resolved on the
           capillary  column,  and if not,  must be reported  as isomeric  pairs.

2.    SUMMARY  OF METHOD

      2.1   Highly volatile organic compounds with low water  solubility are
           extracted  (purged)  from the  sample matrix by  bubbling an inert gas
           through a  5 ml aqueous sample.   Purged sample components are  trapped
           in  a  tube  containing suitable  sorbent materials.  When purging is
           complete,  the sorbent tube is  heated and  backflushed  with helium to
           thermally  desorb trapped sample components  onto a capillary gas

                                    502.2-3

-------
          chromatography  (GC)  column.   The  column  is  temperature programmed to
          separate  the  method  analytes  which  are then detected with a
          photoionization  detector  (PID)  and  an electrolytic conductivity
          detector  (ELCD)  placed  in  series.   Analytes are-quantitated by.
          procedural  standard  calibration (Sect.3.14).

     2.2  Identifications  are  made by comparison of_the retention times of
          unknown peaks to the retention  times of  standards analyzed under the
          same conditions  used for samples.   Additional confirmatory
          information can  be gained  by  comparing the  relative response from
          the two detectors.   For absolute  confirmation, a gas chromatography/
          mass spectrometry  (GC/MS)  determination  according to USEPA Method
          524.2 is  recommended.

     2.3  This method requires  the use  of a PID to measure target analytes
          that cannot be measured with  an electrolytic conductivity detector.
          If only halogenated  analytes,.such  as the trihaldmethanes are to be
          measured, a PID  is not needed.

3.   DEFINITIONS

     3.1  INTERNAL  STANDARD — A pure analyte(s) added to a solution in known
        '  amount(s) and used to measure the relative responses of other method
          analytes  and surrogates that  are components of the same solution.
          The internal standard must be an analyte that is not a sample.
          component.

     3.2  SURROGATE ANALYTE — A pure analyte(s),  which is,extremely unlikely
          to be found in any sample, and which is  added to a sample aliquot  in
          known amount(s)  before extraction and is measured with the same
          procedures used to measure other sample  components.   The purpose of
          a surrogate analyte  is to monitor method performance with each
          sample.
     3.3
     3.4
     3.5
LABORATORY DUPLICATES  (LD1 and  LD2) — Two sample aliquots taken in
the analytical laboratory and analyzed separately with identical
procedures.  Analyses  of LD1 and LD2 give a measure of the precision
associated with laboratory procedures, but not with sample
collection, preservation, or storage procedures.

FIELD DUPLICATES  (FD1  and FD2)  — Two separate samples collected at
the same time and place under identical circumstances and treated
exactly the same throughout field and laboratory procedures.
Analyses of FD1 and FD2 give a  measure of the precision associated
with sample collection, preservation and storage, as w,ell as with
laboratory procedures.

LABORATORY REAGENT BLANK (LRB)  -- An aliquot of reagent water that
is treated exactly as  a sample  including exposure to all  glassware,
equipment, solvents, reagents,  internal standards, and surrogates
that are used with other samples.  The LRB is used to determine if
                                   502.2-4

-------
     method analytes or other interferences are present in the laboratory
     environment, the reagents, or the apparatus.

3.6  FIELD REAGENT BLANK (FRB) — Reagent water placed in a sample
     container in the laboratory and treated as a sample in all  respects,
     including exposure to sampling site conditions, storage,
     preservation and all analytical procedures.  The purpose  of the FRB
     is to determine if method analytes or other interferences are
     present in the field environment.

3.7  LABORATORY PERFORMANCE CHECK SOLUTION (LPC) -- A solution of method
     analytes, surrogate compounds, and internal standards used to
     evaluate the performance of the instrument system with respect to a
     defined set of method criteria.

3.8  LABORATORY FORTIFIED BLANK (LFB) — An aliquot of reagent water to
     which known quantities of the method analytes are added in the
    •laboratory.  The LFB is analyzed exactly like a sample, and its
     purpose is to determine whether the methodology is in control, and
     whether the laboratory is capable of making accurate and  precise
     measurements at the required method detection limit.

3.9  LABORATORY FORTIFIED SAMPLE MATRIX (LFM) — An aliquot of an
     environmental sample to which known quantities of the method
     analytes are added in the laboratory.  The LFM is analyzed exactly
     like a sample, and its purpose is to determine whether the sample
     matrix contributes bias to the analytical results.  The background
     concentrations of the analytes in the sample matrix must  be
     determined in a separate aliquot and the measured values  in the LFM
     corrected for background concentrations.

3.10 STOCK STANDARD SOLUTION — A concentrated solution containing a
     single certified standard that is a method analyte, or a
     concentrated solution of a single analyte prepared in the laboratory
     with an assayed reference compound.  Stock standard solutions are
     used to prepare primary dilution standards.

3.11 PRIMARY DILUTION STANDARD SOLUTION — A solution of several analytes
     prepared in the laboratory from stock standard solutions  and diluted
     as needed to prepare calibration solutions and other needed analyte
     solutions.

3.. 12 CALIBRATION STANDARD (CAL) — A solution prepared from the primary
     dilution standard solution and stock standard solutions of the
     internal standards and surrogate analytes.  The CAL solutions are
     used to calibrate the instrument response with respect to analyte
     concentration.
                               502.2-5

-------
     3.13 QUALITY CONTROL SAMPLE (QCS) — A sample matrix containing method
          analytes or a solution of method analytes in a water miscible
          solvent which is used to fortify reagent water or environmental
          samples.  The QCS is obtained from a source external to the
          laboratory, and is used to check laboratory performance with
          externally prepared test materials.

     3.14 PROCEDURAL STANDARD CALIBRATION --  A calibration method where
          aqueous calibration standards are prepared and processed (e.g.
          purged,extracted, and/or derivatized) in exactly the same manner as
          a sample.  All steps in the process from addition of sampling
          preservatives through instrumental analyses are .included in the
          calibration.  Using procedural standard calibration compensates for
          any inefficiencies in the processing procedure.

4.   INTERFERENCES

     4.1  During analysis, major contaminant sources are volatile materials in
          the laboratory and impurities in the inert purging gas and in the
          sorbent trap.  The use of non-polytetrafluoroethylene (PTFE) plastic
          tubing, non-PTFE thread sealants, or flow controllers with rubber
          components in the purging device should be avoided since such
          materials out-gas organic compounds which will be concentrated in
          the trap during the purge operation.  Analyses of laboratory reagent
          blanks (Sect. 9.2) provide information about the presence of
          contaminants.  When potential interfering peaks are noted in
          laboratory reagent blanks, the analyst should change the purge gas
          source and regenerate the molecular sieve purge gas filter.
          Subtracting blank values from sample results is not permitted.

     4.£  Interfering contamination may occur when a sample containing low
          concentrations of volatile organic compounds is analyzed immediately
          after a sample containing relatively high concentrations of volatile
          organic compounds.  A preventive technique is between-sample rinsing
          of the purging apparatus and sample syringes with two portions of
          reagent water.  After analysis of a sample containing high
          concentrations of volatile organic compounds, one or more laboratory
          reagent blanks should be analyzed to check for cross contamination.

     4.3  Special precautions must be taken to analyze for methylene chloride.
          The analytical and sample storage area should be isolated from all
          atmospheric sources of methylene chloride, otherwise random
          background levels will  result.  Since methylene chloride will
          permeate through PTFE tubing, all gas chromatography carrier gas
          lines and purge gas plumbing should be constructed from stainless
          steel or copper tubing.  Laboratory clothing worn by the analyst
          should be clean since clothing previously exposed to methylene
          chloride fumes during common liquid/liquid extraction procedures can
          contribute to sample contamination.

     4.4  When traps containing combinations of silica gel and coconut
          charcoal are used, residual water from previous analyses collects in

                                    502.2-6

-------
          the trap and can be randomly released into the analytical column.
          To minimize the possibility of this occurring, the trap is
          reconditioned after each use as described in Sect. 11.4.

5.   SAFETY

     5.1  The toxicity or carcinogenicity of chemicals used in this method has
          not been precisely defined; each chemical should be treated as a
          potential health hazard, and exposure to these chemicals should be
          minimized.  Each laboratory is responsible for maintaining awareness
          of OSHA regulations regarding safe handling of chemicals used in
          this method.  Additional references to laboratory safety are
          available (5-7) for the information of the analyst.

     5.2  The following method analytes have been tentatively classified as
          known or.suspected human or mammalian carcinogens: benzene, carbon
          tetrachloride, 1,4-dichlorobenzene, 1,2-dichlorethane,
          hexachlorobutadiene, 1,1,2,2-tetrachloroethane,
          1,1,2-trichloroethane, chloroform, 1,2-dibromoethane,
          tetrachloroethene, trichloroethene, and vinyl chloride.  Pure
          standard materjals and stock standard solutions of these compounds
          should be handled in a hood.  A NIOSH/MESA approved toxic gas
          respirator should be worn when the analyst handles high
          concentrations of these toxic compounds.

6.   EQUIPMENT AND SUPPLIES (All specifications are suggested.  Catalog
     numbers are included for illustration only.)

     6.1  SAMPLE CONTAINERS - 40-mL to 120-mL screw cap vials each equipped
          with a PTFE-faced silicone septum.  Prior to use, wash vials and
          septa with detergent and rinse with tap and distilled water.  Allow
          the vials and septa to air dry at room temperature, place in a 105°C
          oven for one hour, then remove and allow to cool in an area known to
          be free of organics.

     6.2  PURGE AND TRAP SYSTEM - The purge and trap system consists of three
          separate pieces of equipment:,  purging device, trap, and desorber.
          Systems are commercially available from several sources that meet
          all of the following specifications.

          6.2.1  The all glass purging device (Figure 1) must be designed to
                 accept 5-mL samples with a water column at least 5. cm deep.
                 Gaseous volumes above the sample must be kept to a minimum
                 (<15 mL) to eliminate dead volume effects.  A glass frit
                .should be installed at the base of the sample chamber so that
                 the purge gas passes through the water column as finely
                 divided bubbles with a diameter of.<3 mm at the origin.
                 Needle spargers may be used, however, the purge gas. must be
                 introduced at a point <5 mm from the base of the water
                 column.
                                    502.2-7

-------
      6.2.2  The trap  (Figure 2) must be at least 25 cm long and have an
             inside diameter of at least 0.105 in.  Starting from the
             inlet, the trap must contain the following amounts of
             adsorbents:  1/3 of 2,6-diphenylene oxide polymer, 1/3 of
             silica gel, and 1/3 of coconut charcoal.  It is recommended
             that 1.0 cm of methyl silicone coated packing be inserted-at
             the inlet to extend the life of the trap.  If it is not
             necessary to analyze for dichlorodifluoromethane,  the
             charcoal  can be eliminated and the polymer increased to fill
             2/3 of the trap.  If only compounds boiling above  35°C are to
             be analyzed, both the silica gel  and charcoal can  be
             eliminated and the polymer increased to fill  the entire trap.
             Before initial  use,  the trap should be conditioned overnight
             at 180°C by backflushing with an  inert gas flow of at least
             20 mL/min. Vent the trap effluent to the room,  not to the
             analytical column.   Prior to daily use,  the trap should be
             conditioned for 10 min at 180°C with backflushing.  The trap
             may be vented to the analytical column during daily
             conditioning;  however,  the column must be run through the
             temperature program prior to analysis of samples.   The use of
             alternative sorbents is acceptable provided the data acquired
             meets  all  quality control  criteria described  in Section 9,
             and provided the purge and desorption procedures specified  in
             Section 11 of the method are not  changed.   Specifically,  the
             purging time,  the purge gas flow  rate,  and the  desorption
             time may  not be changed.   Since many of  the potential
             alternate  sorbents may be  thermally  stable above 180°C,
             alternate  traps may  be desorbed and  baked  out at higher
             temperatures than those described in Section  11.   If higher
             temperatures are used,  the analyst  should  monitor  the  data
             for possible analyte  and/or trap  decomposition.

      6.2.3   The use of the  methyl  silicone  coated  packing is recommended,
             but not mandatory.   The  packing serves a dual purpose  of
             protecting the  adsorbent from aerosols,  and also of insuring
             that the adsorbent is  fully enclosed within the  heated  zone
             of the  trap  thus  eliminating potential cold spots.
             Alternatively,  silanized glass  wool  may  be used  as  a spacer
             at  the  trap  inlet.

      6.2.4   The desorber  (Figure 2) must be capable of rapidly  heating
             the trap to  180°C.  The polymer section of the trap described
             in  Sect.6.2.2 should not be  heated higher  than 200°C or the
             life expectancy  of the trap will  decrease.  Trap failure  is
             characterized by  a pressure drop  in  excess of 3 pounds per
             square  inch  across the trap during purging or by poor
             bromoform  sensitivities.

6.3  GAS CHROMATOGRAPHY SYSTEM

     6.3.1  The GC must be capable of temperature programming and should
            be  equipped with variable-constant differential flow

                               502.2-8

-------
       controllers so that the column flow rate will remain constant
       throughout desorption and temperature program operation.  The
       column oven may need to' be cooled to <10°C (Sect. 6,3.3), and
       therefore, a subambient oven coritroller may be required.

6.3.2  Capillary Gas Chromatography Columns.  Any gas chromatography
       column that meets the performance specifications of this
       method may be used.  Separations of the calibration mixture
       must be equivalent or better than those described in this
       method.  If other GC columns or temperature programs are
       used, or whenever these procedures are changed, the method
       performance data in Sect. 9.3 must be repeated.  Three useful
       columns have been identified: column 1 (Sect. 6.3.3) and
       column 2 (Sect. 6.3.4) both provide satisfied separations for
       sixty organic compounds.  Column 3 (Sect. 6.3.5), which has
       been satisfactorily demonstrated for the GC/MS method 524.2,
       may also be used.

6.3.3  Column 1- 60m long x 0.75mm'ID VOCOL (Supelco, Inc.)
       wide-bore capillary column with 1.5 jim film thickness, or
       equivalent.  The flow rate of helium carrier gas is adjusted
       to about 6 mL/min.  The'column temperature is held for 8 min
       at 10°C, .then programmed to1 180°C at 4°C/min, and held until
       all expected compounds have eluted.  A sample chromatogram
       obtained with this column is presented in Figure 3.
       Retention times that may be anticipated with this column are
       listed in Table 1.  Data obtained with this column is
       presented in Sect. 13 and 17.

6.3.4  Column 2 - 105m long x 0.53mm ID, RTX-502.2  (O.T
       Corporation/T.ESTEK Corporation) mega-bore capillary column,
       with 3.0 im film thickness, or equivalent.  The  flow rate of
       helium carrier gas is adjusted to about 8 mL/min.  The column
       temperature is held for  10 min at 35°C, then programmed to
       200°'C  at 4°C/min, and held until all expected compounds have
       eluted.  A sample chromatogram obtained with this column is
       presented in Figure 4.   Retention times that may be
       anticipated with this column are listed in Table 3.  Data
       obtained with this column is presented in Sect.  13 and 17.

6.3.5  Column 3 - 30 m long x 0.53 mm ID D8-624 mega-bore  (J&W
       Scientific, Inc.) column with 3 p.m film thickness.

6.3.6  A  series configuration of a high temperature photoionization
       detector  (PID) equipped  with 10 eV (nominal) lamp and
       electroconductivity detector  (ELCD)  is required.  This allows
       the  simultaneous analysis of VOCs that are aromatic or
       unsaturated by photoionization detector and  organohalide by
       an electrolytic conductivity detector.

6.3.7  A  Tracor  703 photoionization detector and a  Tracer Hall model
       700-A  detector connected  in  series' with a short  piece of

                          502.2-9

-------
             uncoated  capillary  tube, 0.32 mm  ID was used to develop the
             single  laboratory method performance data described  in
             Sect.13.   The  system  and operating conditions used to collect
             these data are as follows:
            Column:
            The purge-and-trap  Unit:
            PID detector  base temperature:
            Reactor tube:
            Reactor temperature:
            Reactor base  temperature:
            Electrolyte:
            Electrolyte flow rate:
            Reaction gas:
            Carrier gas plus make-up gas:
Column 1 (Sect.6.3.3)
Tekmar LSC-2
250°C
Nickel 1/16 in. OD
810°C
250°C
100% n-propyl  alcohol
0.8 mL/min
Hydrogen at 40 mL/min
Helium at 30 mL/min
     6.3.8  An O.I. Model 4430 photoionization detector mounting together
            with the model 4420 electrolytic conductivity detector  (ELCD)
            as a dual detector set was used to develop the single
            laboratory method performance data for column 2 described in
            Sect.  13.  The system and the operating conditions used to
            collect these data are as follows:
            Column:
            The purge-and-trap unit:
            Reactor tube:

            Reactor temperature:
            Reactor base temperature:
            Electrolyte:
            Electrolyte flow rate:
            Reaction gas:
            Carrier gas plus make-up gas:
Column 2 (Sect.6.3.4)
O.I. 4460A
Nickel 1/16 in. OD
   & .02in.ID
950°C
250°C
100 % n-propyl  alcohol
0.050 mL/min
Hydrogen at 100 mL/min
Helium at 30 mL/min
6.4  SYRINGE AND SYRINGE VALVES

     6.4.1  Two 5-mL glass hypodermic syringes with Luer-Lok tip.

     6.4.2  Three 2-way syringe valves with Luer ends.

     6.4.3  One 25-/iL micro syringe with a 2 in x 0.006 in ID, 22° bevel
            needle (Hamilton #702N or equivalent).

     6.4.4  Micro syringes - 10, 100 /zL.

     6.4.5  Syringes - 0.5, 1.0, and 5-mL, gas tight with shut-off valve.

6.5  MISCELLANEOUS

     6.5.1  Standard solution storage containers - 15-mL bottles with
            PTFE-lined screw caps.
                              502.2-10

-------
7.   REAGENT AND STANDARDS

     7.1  TRAP PACKING MATERIALS

          7.1.1  2,6-Diphenylene oxide polymer, 60/80 mesh, chromatographic
                 grade (Tenax GC or equivalent).

          7.1.2  Methyl silicone packing (optional) - OV-1 (3%) on
                 Chromosorb-W, 60/80 mesh or equivalent.

          7.1.3  Silica gel - 35/60 mesh, Davison, grade 15 or equivalent.

          7.1.4  Coconut charcoal - Prepare from Barnebey Cheney, CA-580-26
                 lot #M-2649  (or equivalent) by crushing through 26 mesh
                 screen.

     7.2  REAGENTS

          7.2.1  Ascorbic  acid - ACS Reagent grade, granular.

          7.2.2  Sodium thiosulfate - ACS Reagent grade, granular.

          7.2.3  Hydrochloric acid  (1+1) - Carefully add a measured volume of
                 cone. HC1 to equal volume of  reagent water.

          7.2.4  Reagent water - It should be  demonstrated to  be free of
                 interferences.  Prepare reagent water  by passing tap water
                 through a filter bed containing about  0.5 kg  of activated
                 carbon, by using a water purification  system, or by boiling
                 distilled water for 15 min followed by a 1-h  purge with  inert
                 gas while the water temperature is held at 90°C.  Store  in
                 clean, narrow-mouth bottles with PTFE-lined septa and  screw
                 caps.

          7.2.5  Methanol  - demonstrated to be free of  analytes.

          7.2.6  Vinyl chloride  - 99.9% pure vinyl chloride is available  from
                 Ideal Gas Products, Inc., Edison, New  Jersey  and from
                 Matheson, East  Rutherford, New Jersey.  Certified mixtures of
                 vinyl chloride  in  nitrogen at 1.0 and  10.0 ppm  (v/v) are
                 available from  several sources.

     7.3  STOCK  STANDARD SOLUTIONS  - These solutions may be purchased as
          certified  solutions or prepared from pure standard materials  using
          the  following procedures:

          7.3.1  Place about  9.8 mL of methanol  into a  10-mL ground-glass
                 stoppered volumetric flask.   Allow the flask  to  stand,
                 unstoppered, for about  10 min or  until all alcohol-wetted
                 surfaces  have dried.  Weigh to the nearest 0.1 mg.
                                    502.2-11


-------
           7.3.2  If the analyte is a liquid at room temperature, use a 100-fj.l
                  syringe and immediately add two or more drops of reference
                  standard to the flask.  Be sure that the reference standard
                  falls directly into the alcohol without contacting the neck
                  of the flask.  If the analyte is a gas at room temperature,
                  fill a 5-mL valved gas-tight syringe with the standard to the
                  5.0-mL mark, lower the needle to 5 mm above the methanol
                  meniscus, and slowly inject the standard into the neck area
                  of the flask.  The gas will rapidly dissolve in the methanol.

           7.3.3  Reweigh,  dilute to volume, stopper, then mix by inverting the
                  flask several times.  Calculate the concentration in
                  micrograms per microliter from the net gain in weight.  When
                  compound  purity is certified at 96% or greater, the weight
                  can be used without correction to calculate the concentration
                  of the stock standard.

           7.3.4  Store stock standard solutions in 15-mL bottles equipped with
                  PTFE-lined screw caps.   Methanol  solutions  prepared from
                  liquid analytes are stable for at least four weeks when
                  stored at 4°C.   Methanol  solutions prepared from gaseous
                  analytes  are not stable  for more  than one week when stored at
                  <0°C;  at  room temperature, they must  be discarded after  one
                  day.   Storage time may be extended only if  the analyte proves
                  their validity  by analyzing quality control  samples.

      7.4   PRIMARY DILUTION STANDARD SOLUTION -  Use stock standard solutions to
           prepare primary  dilution standard solutions  that contain the
           analytes  in  methanol.   The  primary dilution  standards should  be
           prepay 'd  at  concentrations  that  can be easily diluted to pr epare
           aqueous calibration standard  solutions  (Sect.  9.2)  that will  bracket
           the  working  concentration range.   Store  the  primary dilution
           standard  solutions  with  minimal  headspace and check frequently  for
           signs  of  deterioration  or evaporation, especially  just  before
           preparing calibration  standard solutions  from them.   Storage  times
           described for  stock standard  solutions in Sect.  7.3.4 also  apply to
           primary dilution  standard solutions.

      7.5   INTERNAL  STANDARD SOLUTION  -  Prepare  a fortified solution  containing
           l-chloro-2-fluorobenze or fluorobenzene  and  2-bromo-l-chloropropane
           in methanol using the procedures  described in  Sect.  7.3  and 7.4.   It
           is recommended that the  primary dilution  standard  be  prepared at  a
           concentration of 5  jug/mL  of each  internal  standard  compound.  The
           addition  of 10 /zL of such a standard  to  5.0  mL of  sample or
           calibration standard would be equivalent  to  10 M9/L.

8.   SAMPLE COLLECTION. PRESERVATION. AND STORAGE

     8.1   SAMPLE  COLLECTION AND DECHLORINATION

          8.1.1   Collect all  samples in duplicate.   If  samples,.such as
                  finished drinking water, are suspected to contain residual

                                   502.2-12

-------
8.1.2
            chlorine,  add a dechlorinating agent the bottle.  The
            preferred  dechlorinating agent is sodium thiosulfate,  but
            ascorbic acid may also be used.   Add 3 mg of sodium
            thiosulfate or 25 mg of ascorbic acid per 40 ml of sample to
            the sample bottle before filling  NOTE: If the residual
            chlorine is likely to be present > 5 mg/L, a determination of
            the amount of the chlorine may be necessary.  Diethyl-p-
            phenylenediamine (DPD) test kits are commercially available
            to determine residual chlorine in the field.  Add an addi-
            tional 3 mg of sodium thiosulfate or 25 mg of ascorbic acid
            per each 5 mg/L of residual chlorine.

            When sampling from a water tap,  open the tap and allow the
            system to  flush until the water temperature has stabilized
            (usually about 10 min). • Adjust the flow to about 500 mL/min
            and collect duplicate samples containing the desired dechlo-
            rinating agent from the flowing stream.

     8.1.3  When sampling from an open body of water, partially fill a
            1-quart wide-mouth bottle or 1-L beaker with sample from a
            representative area.  Fill duplicate sample bottles contain-
            ing the desired dechlorinating.agent with sample from the
            larger container.

     8.1.4  Fill sample bottles to overflowing, but take care not to
            flush out  the rapidly dissolving dechlorinating agent.  No
            air bubbles should pass through the sample as the bottle is
            filled,.or be trapped ,in the sample when the bottle is
            sealed.

8.2 SAMPLE PRESERVATION

     8.2.1  Adjust the pH of all samples to < 2 at the time of collect-
            ion, but after dechlorination, by carefully adding two drops
            of  1:1 HC1 for each 40 mL of sample.  Seal the sample bot-
            tles, Teflon face down, and mix for 1 min.  Exceptions to the
            acidification requirement are detailed in Sections 8.2.2 and
            8.2.3. NOTE: Do not mix the ascorbic acid or sodium thiosul-
            fate with the HC1 in the sample bottle prior to sampling.

     8.2.2  When sampling for THM analysis only, acidification may be
            omitted if sodium thiosulfate is used to dechlorinate the
            sample.  This exception to acidification does not apply if
            ascorbic acid is used for dechlorination.

     8.2.3  If  a sample foams vigorously when HC1  is added, discard that
            sample.  Collect a set of duplicate samples but do not acidi-
            fy  them.  These samples must be flagged as  "not acidified"
            and must be stored at 4°C or below.  These  samples must be
            analyzed within 24 hr of collection time if they are to be
            analyzed for any compounds other than THMs.


                              502.2-13


-------
           8.2.4  The samples must be chilled to about 4°C when collected and
                  maintained at that temperature until analysis.  Field samples
                  that will not be received at the laboratory on the-day of
                  collection must be packaged for shipment with sufficient ice
                  to ensure that they will arrive at the laboratory with a
                  substantial amount of ice remaining in the cooler.

      8.3  SAMPLE STORAGE

           8.3.1  Store samples at < 4°C  until  analysis.   The  sample storage
                  area must be free of organic solvent vapors  and direct or
                  intense light.

           8.3.2  Analyze all  samples within 14 days of collection.   Samples
                  not analyzed within this period must be discarded  and re-
                  placed.

      8.4  FIELD REAGENT BLANKS (FRB)
           8.4.1
          8.4.2
        Duplicate FRBs must be handled along with each sample set,
        which is composed of the samples collected from the same
        general  sample site at approximately the same time.  At the
        laboratory,  fill  field blank sample bottles with reagent
        water and sample  preservatives,  seal,  and ship to the sam-
        pling site along  with empty sample bottles and back to the
        laboratory with filled sample bottles.   Wherever a set of
        samples  is shipped and stored,  it is accompanied by appropri-
        ate blanks.   FRBs must remain hermetically sealed until
        analysis.

        Use the  same procedures used for samples to add sodium thio-
        sulfate  or ascorbic acid and HC1  to blanks (Sect.  8.1.1).
        The same batch of ascorbic  acid  and HC1  should be used for
        the field  reagent blanks in  the  field.
9.   QUALITY CONTROL
     9.1
     9.2
Quality control  (QC)  requirements  are  the  initial  demonstration of
laboratory capability followed  by  regular  analyses  of  laboratory
reagent blanks,  field reagent blanks,  and  laboratory fortified
blanks.  A method detection  limit  (MDL) must also  be determined for
each analyte.  The laboratory must maintain records to document the
quality of the data generated.  Additional  quality'control practices
are recommended.

Initial demonstration  of low system background.  Before any samples
are analyzed, it must  be demonstrated  that a laboratory reagent
blank (LRB) is reasonably free  of contamination that would prevent
the determination of  any analyte of concern.  Sources of background
contamination are glassware, purge gas, sorbents, and equipment.
Background contamination must be reduced to an acceptable level
                                   502.2-14

-------
     before proceedi'ng with the next section.  In general background from
     method analytes should be below the method detection limit.

9.3  Initial demonstration of capability.

     9.3.1  Demonstration of laboratory accuracy and precision.  Analyze
            four to seven replicates of a laboratory fortified blank
            containing each analyte of concern at a concentration in the
            range of 0.1-5 /zg/L.  This concentration should represent a
            concentration of ten times the MDL or a concentration hear
            the middle of the calibration range demonstrated (Sect,. 10).
            It is recommended that a QCS from a source different than the
            calibration standards be used for this set of LFBs, since it
            will serve as a che.ck to verify the accuracy of the standards
            used to generate the calibration curve.  This is particularly
            useful if the laboratory is using-'the method for the first
            time, and has no historical data base for standards.  Prepare
            each replicate by adding an appropriate aliquot of a quality
            control sample to reagent water.  Also add the appropriate
            amounts of internal standard and surrogates if they are being
            used.  If it is expected that field samples will contain a
            dechlorinating agent and HC1, then add these to the LFBs in
            the same amounts prescribed in Sect. 8.1.1.  If only THMs are
            to be determined and field samples do not contain HC1, then
            do not acidify LFBs.  Analyze each replicate according to the
            procedures described in Sect. 11.

     9.3.2  Calculate the measured concentration of each analyte in each
            replicate, the mean concentration of each analyte in all
            replicates, and mean accuracy (as mean percentage of true
            value) for each analyte, and the precision (as relative
            standard deviation, RSD) of the measurements for each analy-
            te.

     9.3.3  For each analyte and surrogate, the mean accuracy, expressed
            as a percentage of the true value, should be 80-120% and the
            RSD should be <20%.  Some analytes, particularly the early
            eluting gases and late eluting higher molecular weight com-
            pounds, are measured with less accuracy and precision than
            other analytes.  If these criteria are not met for an analy-
            te, take remedial action and repeat the measurements for that
            analyte to demonstrate acceptable performance before samples
            are analyzed.

     9.3.4  To determine the MDL, analyze a minimum of 7 LFBs prepared at
            a low concentration.  MDLs in Tables 2 and 4 were calculated
            from samples fortified at 0.1 /zg/L, which can be used as a
            guide, or use calibration data to estimate a concentration
            for each analyte that will yield a peak with a 3-5 signal to
            noise ratio.  Analyze the 7 replicates as described in
            Sect.11, and on a schedule that results in the analyses being
            conducted over several days.  Calculate the mean accuracy and

                              502.2-15

-------
             standard deviation for each analyte.   Calculate the MDL using
             procedures described in Ref.  4.   The  equation for this  calcu-
             lation is also in Sect. 13.3.

      9.3.5  Develop and maintain a system of control  charts to plot the
             precision and accuracy of analyte and surrogate measurements
             as a function of time.  Charting of surrogate recoveries is
             an especially valuable activity  since these are present in
             every sample and the analytical  results will  form a signifi-
             cant record of data quality.

 9.4   Laboratory reagent blanks (LRBs).   With each batch of samples
      processed as a group within a work shift,  analyze  a laboratory
      reagent blank to determine the background  system contamination.
      LRBs  should contain the same additives  (dechlorinating agent and
      HC1)  as field samples.

 9.5   Assessing Laboratory .Performance.   With each batch of samples
      processed as a group within a work shift,  analyze  a  single laborato-
      ry  fortified blank (LFB)  containing  each analyte of concern at a
      concentration as determined in 9.3.1.   LFBs  should contain a dechlo-
      rination  agent and/or HC1  as appropriate to  match  the field samples
      being  analyzed.   The minimum frequency  of  LFB analysis is  once every
      twelve hours.   Use the  criteria described  in 9.3.3 to evaluate the
      accuracy  of the  measurements,  and  to  estimate whether the  method
      detection limits can be obtained.   If acceptable accuracy  and  method
      detection limits cannot be achieved,  the problem must be  located and
      corrected before further  samples are  analyzed.   Data  from  all  field
      samples analyzed since  the last acceptable LFB should be  considered
      suspect,  and duplicate  samples should be analyzed,  if they are
      available,  after the problem has been corrected.   LFB results  should
     'be  added  to  the  on-going  control charts  to document data  quality.

      Since  the calibration check sample  in Sect.  10.3.2 and the LFB are
      made the  same  way and since procedural  standards are  used, the
      sample  analyzed  here may  also  be used as the calibration  check in
      Sect.  10.3.2.

9.6  Assessing  the  Internal  Standard.   If  internal standard  calibration
      is  used,  the analyst must  assess the  response of the  internal
      standard  in  every LRB,  FRB,  LFB, CAL, and field  sample.  The IS
      response  (peak height or  peak  area units) must be  within   20% of
     the mean  peak  response  of  the  IS in the  CAL  standards  used to
     develop the  calibration.   If  this criteria cannot  be  met,  take
     remedial  action.   If there  are  interferences in  field  samples  that
     affect the measurement  of  the  internal  standard,  external  standard
     calibration  should  be used  (Sect. 10.3.2).

9.7  Assessing  the  Surrogate Analyte.  Calculate  the amount of  the
     surrogate  analyte  recovered  in  each LRB, LFB, FRB,  CAL, and field
     sample  (Sect.10).   If the  surrogate recovery in blanks or  calibra-
     tion standards does  not meet the criteria in Sect.  9.3.3., take

                              502.2-16

-------
          remedial  action.   If the  surrogate  recovery  in  a  field  sample does
          not  meet  the  criteria in  Sect  9.3.3.,  and  data  from  LFBs  shows  the
          laboratory  to be  in  control, reanalyze the sample.

     9.8   If a water  sample is contaminated with an  analyte, verify that  it is
          not  a sampling error by analyzing a field  reagent blank.   The
          results  of  these  analyses will  help define contamination  resulting
          from field  sampling, storage and transportation activities.   If the
          field reagent blank  shows unacceptable contamination, the analyst
          should identify and  eliminate  the contamination.

     9.9   At least  quarterly,  replicates of laboratory fortified  blanks should
          be evaluated  to determine the  precision of the  laboratory measure-
          ments.  Add these results to the on-going  control charts  to  document
          data quality.

     9.10 At least  quarterly,  analyze a  quality control  sample (QCS) from an
          external  source.   If measured  analyte concentrations are  not of
          acceptable  accuracy, check the entire analytical  procedure to locate
          and correct the problem source.

     9.11 Sample matrix effects have not been observed when this  method is
          used with distilled  water, reagent  water,  drinking water, and ground
          water.  Therefore, analysis of a laboratory  fortified sample matrix
          (LFM) is not  required.

     9.12 Numerous other quality control measures are  incorporated  into   other
          parts of this procedure,  and  serve  to alert  the analyst to  poten-
          tial problems.

10.   CALIBRATION AND STANDARDIZATION

     10.1 Demonstration and documentation of  acceptable initial calibration  is
          required before any samples are analyzed.   In addition, acceptable
          performance must be confirmed  intermittently throughout analysis of
          samples by performing continuing  calibration checks.  These checks
          are required at the beginning  of each work shift, but no  less than
          every 12 hours.  Additional periodic calibration  checks are good
          laboratory practice.  Since this  method uses procedural standards,
          the analysis of the laboratory fortified blank, which is  required  in
          Sect. 9.5,  may be used here as a calibration check sample.

     10.2 PREPARATION OF CALIBRATION STANDARDS

          10.2.1 The number of calibration solutions (CALs) needed depends on
                 the calibration range desired.  A minimum of three CAL solu-
                 tions is required to calibrate a range of a factor of 20 in
                 concentration.  For a factor of 50 use at least four stan-
                 dards, and for a  factor of  100 at least five standards.   One
                 calibration  standard should contain each  analyte of concern
                 at a  concentration 2 to 10  times greater  than the method
                 detection limit (Table 2 and 4) for that  compound.  The  other

                                   502.2-17

-------
            CAL standards should contain each analyte of concern at
            concentrations that define the range of the sample analyte
            concentrations.  When internal standard calibration is being
            used, every CAL solution contains the internal standard at
            same concentration (10 /Kj/L).

     10.2.2 To prepare a calibration standard, add an appropriate volume
            of a primary dilution standard solution to an aliquot of
            reagent water in a volumetric container or sample syringe.
            The reagent water used should also contain the appropriate
            dechlorinating agent and/or HC1 so as to match the field
            samples to be analyzed.   Use a microsyringe and rapidly
            inject the alcoholic standard into the water.  Remove the
            needle as quickly as possible after injection.  Accurate
            calibration standards can be prepared by injecting 20 fiL of
            the primary dilution standards to 25 mL or more of reagent
            water using the syringe described in section 6.4.3.  Aqueous
            standards are not stable in volumetric container and should
            be discarded after one hour unless transferred to sample
            bottle and sealed immediately as described in Sect. 8.1.2.

10.3 CALIBRATION

     10.3.1 External  standard calibration.  Starting with the standard of
            lowest concentration, analyze each calibration standard
            according to Sect.  11 and tabulate peak height or area re-
            sponse versus the concentration in the standard.   The results
            can be used to prepare a calibration curve for each compound.
            Alternatively, if the ratio of response to concentration
            (calibration factor)  is  a constant over the working range
            (<10% relative standard  deviation),  linearity through the
            origin can be assumed and the average ratio or calibration
            factor can be used  in place of a calibration curve.

     10.3.2 Internal  standard calibration.  The organohalides  recommended
            as internal  standards are:  l-chloro-2-fluorobenze or 2-brom-
            o-1-chloropropane and fluorobenzene.   The internal  standard
            is added  to the sample just before purging.   Check the valid-
            ity of the internal  standard response factors daily by ana-
            lyzing a  calibration  standard.   NOTE:  Since the calculated
            concentrations can  be strongly biased by inaccurate detector
            response  measurements for the internal  standard or by coelut-
            ion of an unknown with the  internal  standard,  it  is required
            that the  area measurement of the internal  standard of each
            sample be within  ±  3  standard deviations of those obtained
            from calibration  standards,  or ± 20%  of the mean  response ob-
            tained from calibration  standards,  whichever is greater.   If
            they do not,  then internal  standards  can not be used.

     10.3.3 Following analysis,  tabulate peak height or area  responses
            against concentration for each  compound  and the internal
                              502.2-18

-------
             standard.   Calculate the response factor (RF) for each com-
             pound using Equation 1.

             Equation 1

             RF = iAsl_LC,. 1
                      (O (C.)

             where

             As   = Response for the analyte to be measured
             Ais = Response for the internal standard
             Cjs = Concentration of the internal  standard (#g/L)
             Cs   = Concentration  of the analyte to be measured  (/ag/L)

             If RF value over  the working  range  is constant  (<  10%  RSD),
             the average RF can be used for calculations.  Alternatively,
             the results can be used  to plot a calibration curve  of re-
             sponse  versus analyte ratios, A0/A,   vs.  C /C- .
                                           S  IS      S'   1S

     10.3.4  The working calibration  curve or calibration  factor  must be
             verified by the measurement of one or more  calibration  stan-
             dards.   This must  be  done at  least once  each work  shift, but
             no  less  than once  every  twelve hours.  Additional  periodic
             calibration checks are good laboratory practice.   It is
             highly  recommended that  an additional calibration  check be
             performed at the end  of  any cycle of continuous  instrument.
             operation,  so that each  set of field samples  is  bracketed by
             calibration  check standards.  It is also recommended that
             more that one concentration of continuing calibration  stan-
             dard be analyzed, in  order to evaluate the  accuracy of the
             calibration  at more than one point.   If the response for any
             analyte varies from the  predicted response  by more than ±
             20%, the test must be repeated using a fresh calibration
             standard.   If the results still do not agree, generate a new
             calibration curve.  Any  field samples analyzed since the last
             acceptable  calibration check should be considered suspect,
             and should  be reanalyzed if they are available.

10.4 CALIBRATION FOR VINYL CHLORIDE USING A CERTIFIED GASEOUS MIXTURE
     (OPTIONAL)

     10.4.1 Fill the purging  device with  5.0  mL  of reagent water or
            aqueous  calibration standard,  and  add internal standards.

     10.4.2 Start to purge the aqueous mixture (Sect. 7.2.6). Inject a
            known volume (between 100 and  2000 0L)  of the calibration gas
            (at room temperature) directly into  the  purging  device  with a
            gas tight syringe.   Slowly inject  the gaseous sample through
            the aqueous  sample inlet  needle.   After  completion, inject  2
            mL of clean  room  air  to  sweep  the  gases  from the inlet  needle


                              502.2-19

-------
                 into the purging device. Inject the gaseous standard before
                 five min of the 11-min purge time have elapsed.

          10.4.3 Determine the aqueous equivalent concentration of vinyl
                 chloride standard injected in jug/L, according to the  equa-
                 tion:
                         S = 0.51 (C) (V)

                 where:  S
                         Equation 1
Aqueous equivalent concentration of vinyl
chloride standard in jug/L;
Concentration of gaseous standard in ppm (v/v);
Volume of standard injected in milliliter
                         C
                         V

11.  PROCEDURE
                                                                i.-'
     11.1 INITIAL CONDITIONS

          11.1.1 Recommended chromatographic conditions are summarized in
                 Sect. 6.3.  Other columns or GC conditions may be used if the
                 requirements of Sect. 9.3 are met.

          11.1.2 Calibrate the system daily as described in Sect.  10.

          11.1.3 Adjust the purge gas (nitrogen or helium) flow rate to 40 mL-
                 /min.  Attach the trap inlet to the purging device and open
                 the syringe valve on the purging device.

     11.2 SAMPLE INTRODUCTION AND PURGING

          11.2.1 To generate accurate data, samples and calibration standards
                 must be analyzed under identical conditions.  Remove the
                 plungers from two 5-mL syringes and attach a closed syringe
                 valve to each.  Allow the sample to come to room temperature,
                 open the sample (or standard) bottle, and carefully pOur the
                 sample into one of the syringe barrels to just short of
                 overflowing.  Replace the syringe plunger, invert the sy-
                 ringe, and compress the sample.  Open the syringe valve and
                 vent any residual  air while adjusting the sample volume to
                 5.0 mL.  Add 10 pi of the internal calibration standard to
                 the sample through the syringe valve. Close the valve.  Fill
                 the second syringe in an identical manner from the same
                 sample bottle.  Reserve this second syringe for a reanalysis
                 if necessary.

          11.2.2 Attach the sample syringe valve to the syringe valve on the
                 purging device.  Be sure that the trap is cooler than 25°C,
                 then open the sample syringe valve and inject the sample into
                 the purging chamber.  Close both valves and initiate purging.
                 Purge the sample for 11.0 ± 0.1 min at ambient temperature.
                 Note: Ambient room temperature must be relatively constant.
                 If it varies by more than 10°C during an analysis day, or be-

                                   502.2-20

-------
                 tween calibration and sample analysis,  precision and accuracy
                 of some analytes will be affected.

     11.3 SAMPLE DESORPTION - After the 11-min purge,  couple the trap to the
          chromatograph by switching the purge and trap  system to the desorb
          mode,  initiate the temperature program sequence of the gas chromato-
          graph  and start data acquisition.   Introduce the trapped materials
          to the GC column by rapidly heating the trap to 180°C while backflu-
          shing  the trap with an appropriate inert gas flow for 4.0 ±0.1 min.
          While  the extracted sample is being introduced into the gas chro-
          matograph, empty the purging device using the  sample syringe and
          wash the chamber with two 5-mL flushes of reagent water.

     11.4 TRAP RECONDITIONING - After desorbing the sample for four min,
          recondition the trap by returning  the purge  and trap system to the
          purge  mode.  Maintain the trap temperature at  180°C.  After approxi-
          mately seven min, turn off the trap heater and open the syringe
          valve  to stop the gas flow through the trap.  When the trap is cool,
          the next sample can be analyzed.

12.   DATA ANALYSIS AND CALCULATIONS

     12.1 Identify each analyte in the sample chromatogram by comparing the
          retention time of the suspect peak to retention times generated by
          the calibration standards, the LFB and other fortified quality
          control samples.  If the retention time of the suspect peak agrees
          within ± 3 standard deviations of  the retention times of those
          generated by known standards (Table 1 and 3) then the identification
          may be considered as positive.  If the suspect peak falls outside
          this range or coelutes with other  compounds  (Table 1 and 3), then
          the sample should be reanalyzed.  When applicable, determine the
          relative response of the alternate detector  to the analyte.  The
          relative response should agree to  within- 20% of the relative re-
          sponse determined from standards.

     12.2 Xylenes and other structural isomers can be  explicitly identified
          only if they have sufficiently different GC  retention times. Accept-
          able resolution is achieved if the height of the valley between two
          isomer peaks is less than 25% of the sum of  the two peak heights.
          Otherwise, structural isomers are  identified as isomeric pairs.

     12.3 When both detectors respond to an  analyte, quantitation is usually
          performed on the detector which exhibits the greater response.
          However, in cases where greater specificity  or precision would
          result, the analyst may choose the alternate detector.  Do not
          extrapolate beyond the calibration range established in Sect. 10.
          If peak response exceeds the highest calibration standard, a dupli-
          cate sample must be diluted and reanalyzed.   Use only the multi-
          point  calibration data obtained in Sect. 10  for all calculations.
          Do not use the daily calibration verification standard to quantitate
          method analyte in samples.


                                   502.2-21

-------
      12.4 Determine the  concentration  of  the  unknowns  when  external  standards
           are used, by using  the  calibration  curve  or  by  comparing the  peak
           height or area of the unknowns  to the  peak height  or  area  of  the
           standards as follows:

           Concentration  of unknown  (jug/L) = (Peak height  sample/Peak height
           standard) x Concentration of standard  (/zg/L).

      12.5 Calculate analyte and surrogate concentrations  when internal  stan-
           dards are used with the equation in Sect. 10.3.3.

      12.6 Calculations should utilize  all available digits of precision, but
           final  reported concentrations should be rounded to an appropriate
           number of significant figures(one digit of uncertainty).   Experience
           indicates that three significant figures may be used for concentra-
           tions  above 99 jug/L, two significant figures for concentrations
           between 1 to 99 pg/L, and 1  significant figure  for lower concentra-
           tions.

      12.7 Calculate the total  trihalomethane concentrations by summing the
           four individual trihalomethane concentrations in /zg/L.

 13.   METHOD  PERFORMANCE

      13.1 This method  was tested in a single laboratory using reagent water
           fortified at 10 0g/L (1).   Single laboratory precision and accuracy
           data for  each detector are presented for the method analytes in
           Tables 2  and 4.

      13.2 Method detection  limits  for these analytes have  been  calculated from
           data collected  by  forcifying  reagent water at 0.1  /ug/L.(l).  These
           data are  presented  in Tables  2  and  4.
      13.3 Method  detection  limits  were calculated using the formula:

          MDL = S t,

          where:
LCn-1,1-alpha = 0.99)
          t
-------
          preparing standards and sample preservatives.  All are used in
          extremely small amounts and pose no threat to the environment.

15.   WASTE MANAGEMENT

     15.1 There are no waste management issues involved with this method.  Due
          to the nature of this method, the discarded samples are chemically
          less contaminated than when they were collected.

16.   REFERENCES

     1.    Ho, J.S., A Sequential Analysis for Volatile Organics in Water by
          Purge and Trap Capillary Column Gas Chromatograph with Photoionizat-
          ion and Electrolytic Conductivity Detectors in Series, Journal of
          Chromatographic Science 27(2) 91-98, February 1989.

     2.    Kingsley, B.A., Gin, C., Coulson, D.M., and Thomas, R.F., Gas
          Chromatographic Analysis of Purgeable Halocarbon and Aromatic
          Compounds in Drinking Water Using Two Detectors in Series, Water
          Chlorination, Environmental Impact and Health Effects, Volume 4, Ann
          Arbor Science.

     3.    Bellar, T.A., and J.J. Lichtenberg, The Determination of Halogenated
          Chemicals in Water by the Purge and Trap Method, Method 502.1, U.S.
          Environmental Protection Agency, Environmental Monitoring and
          Support Laboratory, Cincinnati, Ohio 45268, April, 1981.

     4.    Glaser, J.A., D.L. Foerst,  G.D. McKee,  S.A. Quave, and W.L. Budde,
          Trace Analyses for Wastewaters, Environ. Sci. Technol., 15, 1426,
          1981.

     5.    Carcinogens - Working with  Carcinogens, Department of Health,
          Education,  and Welfare, Public Health Service, Center for Disease
          Control,  National  Institute for. Occupational  Safety and Health,
          Publication No. 77-206, August, 1977.

     6.    OSHA Safety and Health Standards, (29 CFR 1910), Occupational Safety
          and Health  Administration,  OSHA 2206.

     7.    Safety in Academic Chemistry Laboratories, American Chemical  Society
          Publication, Committee on Chemical  Safety, 4th Edition, 1985.

     8.    Bellar, T.A. and J.J.  Lichtenberg,  The Determination of Synthetic
          Organic Compounds  in Water  by Purge and Sequential Trapping Capil-
          lary Column Gas Chromatography, U.S. Environmental Protection
          Agency, Environmental  Monitoring and Support Laboratory, Cincinnati,
          Ohio, 45268.

     9.    Slater, R.W., Graves,  R.L.  and McKee, G.D., "A Comparison of Preser-
          vation Techniques  for Volatile Organic Compounds in Chlorinated Tap
          Waters,"  U.S. Environmental Protection  Agency, Environmental  Moni-
          toring and  Support Laboratory, Cincinnati, Ohio  45268.
                                   502.2-23

-------
17.  TABLES. DIAGRAMS. FLOWCHARTS. AND VALIDATION DATA

           TABLE 1.  RETENTION TIMES FOR VOLATILE ORGANIC COMPOUNDS
                     ON PHOTOIONIZATION DETECTOR (PID) AND ELECTROLYTIC
                     CONDUCTIVITY DETECTOR (ELCD) FOR COLUMN 1
        Ana1vte(b)
   Retention Time (min)a
PID	ELCD
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23

24

25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
Di chl orodi f 1 uoromethane
Chloromethane
Vinyl Chloride
Bromomethan
Chloroethane
Tri chl orof 1 uoromethane
1,1-Dichl oroethene
Methyl ene Chloride
trans-1 , 2-Di chl oroethene
1, 1-Di chl oroethane
2, 2-Di chl oropropane
cis-1, 2-Di chl oroethene
Chloroform
Bromochl oromethane
1,1, 1-Tri chl oroethane
1 , 1-Di chl oropropene
Carbon Tetrachloride
Benzene
1, 2-Di chl oroethane
Trichl oroethene
1 , 2-Di chl oropropane
Bromodi chl oromethane
Dibromomethane
Cis-1, 3-Di chl oropropene
Toluene
Trans-1 , 3-Di chl oropropene
1,1, 2-Tri chl oroethane
Tetrachl oroethene
1,3-Dichl oropropane
Di bromochl oromethane
1,2-Dibromoethane
Chlorobenzene
Ethyl benzene
1,1,1, 2-Tetrachl oroethane
m-Xyl ene
p-Xylene
o-Xylene
Styrene
Isopropyl benzene
Bromoform
1,1,2, 2-Tetrachl oroethane
1, 2, 3-Tri chl oropropane
n-Propyl benzene
-(c)
-
9.88
-
-
-
6.14
• -
19.30
-
-
23.11
-
-
-
25.21
-
26.10
-
27.99
-
- '
-
31.38
31.95
33.01
-
33.88
-
-
-
36.56
36.72
•
36.98
36.98
38.39
38.57
39.58
-
-
-
40.87
8.47
9.47
9.93
11.95
12.37
13.49
16.18
18.39
19.33
20.99
22.88
23.14
23.64
24.16
24.77
25.24
25.47
-
26.27
28.02
28.66
29.43
29.59
31.41
-
33.04
33.21
33.90
34.00
34.73
35.34
36.59
-
36.80
-
-
-
-
-
39.75
40.35
40.81
-
                                   502.2-24

-------
                             TABLE  1  (CONTINUED)
         Ana1vte(b)
Internal Standards
     Fluorobenzene
     2-Bromo-l-chloropropanec
    Retention Time (min)a
 PIP	ELCD
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
Bromobenzene
1,3, 5-Tri methyl benzene
2-Chlorotoluene
4-Chlorotoluene
tert-Butyl benzene
1,2, 4-Tri methyl benzene
sec-Butyl benzene
p- Isopropyl to! uene
1,3-Dichlorobenzene
1,4-Di chlorobenzene
n-Butyl benzene
1,2-Dichlorpbenzene
1 , 2-Di bromo-3-Chl oropropane
1,2, 4-Tri chlorobenzene
Hexachlorobutadiene
Naphthalene
1, 2, 3-Tri chlorobenzene
40.99
41.41
41.41
41.60
42.71
42.92
43.31
43.81
44.08
44.43
45.20
45.71
-
51.43
51.92
52.38
53.34
41.03
-
41.45
41.63

-
-
-
44.11
44.47
-
45.74
48.57
51.46
51.96
-
53.37
26.84
                    33.08
a. Column and analytical conditions are described in Sect. -6.3.
b. Number refers to peaks in Figure 502.2-1.
c. - Dash indicates detector does not respond.
d. Interferes with trans-l,3-dichloropropene and
     1,1,2-trichloroethane on the column. Use with care.
                                   502.2-25


-------
»— 1
sc
1— 1
_J
z
o
&™4
H-
0
LU^
fcs
a t-«
o z
CJ ^^
•Tf •— ^
Lu o
j^ CJ

a a£
Z 0


»»as

4-1
•r«
>
•r—
• \
i *
u
=1
•a s-
c o
0 •«->
0 U
fi>
tu
O -M
•r- cu
•4-» a
4?
o
s-
4-J
O
a>
1—
UJ
^-*
Q ~-
• M
2_ G
••—

• C
-o o
•»-> -I-
CO 4->
ro *—
i— > — •
CU CU
CtL Q

>-^,^-«
cu S— —
CD CU -^
rO > a
S- 0 =i
cu o •—
> eu
CO rH CM
o o o 10 i— i
O O O t— I i— i





!**•» O CT* CM CO '
CM CO CM LO CO





^^
«^ t"*^ *O ^^ l^ ^^
**-^O> CT^ fj> CD CTi
1 t-i

Z
O I—
 o i—
 LU Z
 Q£ UJ
 a. o

  ..S
 >• OS
 O
    co
    a
o o
CO t-t
   CO
LU as
— J O
CO
   *£.
  • —1
CM O

LU
-JDS
CO O

 o
J=
a.
                      •  C
                    •a  o
                     cu  cu
                    OS Q
                     CU  S- .
                     cn cu .
                     s-  o
                     o  u
                           CD
                                                   i   i   i
                                                    I i—I    CM CO r—I t—I              CMCMr—ILOr^COI~~'—llOi—(CO
                                                    >O*—(OOOC3OCOOOCMCSOCDOOOOOC3CDO

                                                 O O O O O O OCOOOCMOOOCDCDOOOOOO
                                                                                                  r*-cvjiOLncoe\jcococofO
                    o o
                      •   •
                    o o
i   i
                                                   I   I
                                        CM CM IO     i—(
                                        o o o     o
                                        o o o
     •CM              LO CM i—(            -CMLO
    QO              OC3O           QOO
     •   •  I    I   I   I    •••III   •   •   •  I    I
    •Z.Q              C3C3O           2TCDO
                                                 «d- t-- CO     CD          Q CD
                                         III   •  •   •  I   •  I   I   I   ••III
                                                 «d- CM CM     •—(          Z •—I
                                                                                     i—i IO i—i          *$• O O
                                                                                       •   •   'III   •  •   •  I   I
                                                                                     CM r—I CM          CM Z «»•
                           cj cn
                              en 01
     Ol— 00    O           Q i—(               CM^-cO           CDQCO
     ocncnioi   i   i    -01   i   i    ICDCDCDI   i   10   -cni   i
                              cu
                              cu
                           I  c
                           cu re
                           c -c
                           ro +->
                          J= O)

                           cu o
                        CU  £ S-
                        c  o o
                        CU  S- r—
                        M  O J= E
                        C r— CJ S-
                        Nrej=
                                              J=CUr— >,CUc:jcS_
                                              4J -Q >>+-> 4->  CU -(->
                                               cui
                                                               _  .   o
                                                              -Q  CU ^4-
           CU
           c
           re
           CL
           o

           Q.
           O  O)
           S-  C        CU  CU  CU
           O  re
          i— -C  CU
          j=: •*->  c
    CU CU  O  CU  ro

 cu  cu QJ co  o -i-> c
 c  =s zs  i   s-  cu re
 ro i— i—  O  O  O J=
-C  O O  E r—  £ -f->
-M -M +->  O  *"  O CU r-~ r— i—
 ajoos-us-E-c-c-ci
 ES-S--QO-QOO(_IU
                                                                                               CU
                                                                                               c
                                                                                               ro
 cu  cu
 N  N
 c  c
 cu  cu
_Q -Q
 o  o
 S-  S-
 o  o
 cu  cu cu cu
 e  c c c
 O  ro ro CU
 S- J= J= J=
 O 4-> -l-» -l->
 3  CU CU CU I
i—  O O O ,
M-  S- S- S-
•i-  O O O
                              CUOOOOO+JCQ  I   OS-S-&-
                              0)
                                                               s-r—r—Q oa  OQI
                       E  E  E E  e  r;  i-i-jj3oooo-c:-c  i  s-  i   s-  i
                       O  O  O O  OCa CM-OCM-O.CM,
                       S-S-S-S-S-  I  CUCUrO^:j=J=J=  I   I   -•!-   "•!-  ^
                       cacacacaca  c <^+-> cj c_j cjc_jc_5cM«^-r-(C3i—(Oi—< ,
                                                             O J= J= J=
                                                             S-  O (J  O
                                                             O **"• T"" T—
                                                            i— Q Q a
                                                            J=   I   I   I
                                                             O i—( CM i—I
 cu
 c
 cu
-C

 CU  O) CU
 o  c c
 S-  ro re
 o  a. o.
r—  O O
J=  S- S-
 u  a. a.
-r-  O O
o  s- s-
 I   O O
                                                  I  -I- -r-
                                                  Q Q
                                                 C  I   I
                                                 rO CM CO
                                                            502.2-26

-------
                                      en











>,
^.
>
>r_
1 t
o
3
"O
0
o

o
•r—
4-3
>^
^_
O

1 '
£T* o
S *>
i G
z
NH
•
r~
O
O
^~^


*
CM ^
UJ °
^™ -
^-1 ,_}
s s
»- £
t—
o
o

o
.c
d.





















































S-
o
o
cu
4^
cu
0





















S-
o
o
CU
cu
Q















































	 1
O
s


• c:
•o o

CO | ^
ra
,_! >
cu cu
o£ rS


>-
CU S-
CT> CU
ra >
S- 0
cu o
> cu

cu cu
cc: a

>•
cu s_
CD CU
ra >
S- 0
cu o
> cu
.
.j—
ro
C










LOCSJ CSJ CM i— I •— 1 **• CO CO CO •
oo o o ooo oooo
00 0 0 000 OOOZ






** CO CO OV CO CO LO •— I ,— 4 CO IO
COCO CO CM CM IO CM CO CM CO LO







LO co co i^^ en en r*^ co CM ^ en
ooicni i en i i icncncnicnooo
i-H i — 1 • i — 1 i-H i — I





CMi — llOLOl — 1 <£> i— 1 i — 1 LOr-H -CXI
OOOOO OOO OOOO
OOOOO OOO O'OZO








LO *3- LO Cn «*• CMOCO OO OO OO CSJ
CO •— I CTl O CM IO CM i— I •— lOi-HCM







co i— i en co oo CM co «*• I-H en 10 «*•
i **"*** <*^ en en en i ^"* *"™* <""i i \ f^ en ***" ^ c * i i
I-H I-H i-H i-H i-H i-H i-H I-H







CU CU
£ C
ro ro
-C -C CU CU
+J 4-> C C CU O)
CU CU O) CU £1 £=
CUCUCU OO MlslrOrO
c c: c: cucu s_ s_ c c -c -sz
rOCU CU CT3 OOCU CUCU-*J4->
O.O- -r-CUCU-r- r- i— C -Q-OCUCU
oo •oc3i- cu ^:^:cu oooo
S-S- (OCUi — O C: OU-C S-S-S-S-
O-O- +->NOi — CU rCrC+J OOOO
O O CU 3 e 4-> <"" M i- i- CU i — i — i — i —
S-S-C-OCUi — OCUC +J4->O JZJZJZJZ
OOCUO^D>, CCU CUCU&- CJUUU
i — i — IMS-i — O-CUCU^J 1 — 1 — O -i— -i— -i— -r-
-C-CCO>>OCi — i— 1 1 i — S-S-S-S-
o u cui — Q.S- cu ra >,CUCMCM^: cut— i— t— 1—
-•- -r- _Q ^: o a..— .c: a. c - *. u c i i i i
O O i — O S- O 5">-4-> O CU i— 1 CM rO CU CO *3- •— I CM
i i >>rca.oo_c:_£:s-s- •> -s-3 - - - -
CM f— 1 SZ X O I— I 4-> Q.Q- >>•— I r— ( 4J i — CM CM i-H •— 1
„ x+j CUOOICUrOI+J x»,cuo r.^x^
CM f-H UJ "T" >-H Q.SI "^ C CO i — I »-H h~ 1 — ' — ' ' — ' *~^ « — '









1 — 1
o
0






IO
CO







<£)
en






CM
CD
o







co
r--.
o







o

i-H

















cu
c
cu
^
4^
cu
o
s_
o
^~
f~
o
• ^
^
1—









CO
ez>
0






LO
CO







IO
en








1









'








1











cu

(O
-C
+->
cu
E
o
s_
o
3
r—
M-
o
s-
o
r— •
-g?
^J
•^
^
h-










«d-
o






CO
CNJ







en
en








1









1








1











cu

(O
Q-
o
s-
O-
o
s-
o
^~
f~
o
•r—
s_
h-
1
CO
*»
CM
n
r-H











I







1








1






LO
o
0








CM
r-H







en
en











cu

cu
N
e
cu
-Q
^—
>1

4-3
cu
E
•r—
S-
I—
1
^J*
**
CM
n
1— 1









^-
O
1 "III
o






en
1 • 1 1 1
1 * 1 1 1
LO







LO
i cn i i i






i — 1 CM CM i — 1 i-H
O O O O O
OOOOO








>st- o oo «*• en
r— C LO O r- 1 O







r-H en en o en
o o en o en
1 — 1 i-H r-H










CU
e
cu
N
C
CU
-Q
^~
>*j Q)
-C T3

cu s-
E 0
•r~ r-~
S- -C CU 0) CU
1— o c c c
1 CU CU CU
LO ^"~ ^~ ^— f~
X ^*) ^*> ^*) ^>
co c x x x
-•1- i i i
—i > 0 E 0.
0 ••
,-H 0)
S-
4-> CU
(& ^E
13 oo
cu -a
•i- S-
4- ro
•4-J C
t- ra
O 4-*
4- oo
00 i —
CU ro
i — C
Q. S-
E CU
ro 4->
00 S=
C
cu •
> -a
cu o
00 J=
4-9
E CU
0 E
S_
4- -a
S-
TD rO
cu -a
C^—
^_
•i- ra
E 4^
S- 00
cu
4J 1 —
T3 C
s-
cu cu
S»- t -*
cu c:


oo >,
C LQ
o
•r™ i^;
4-> CU
rO C
•1— •£-
CU S-
T3 CU
4^
13 0)
s- -o
ro
-o cu
C i-
rO CU
4*^ 3
00
00
cu cu
> -r—
•i- S-
4-> CU
rO >
— 0
cu u
s- cu
C£
•o
c •
ro CU
+J
00 >5
cu ^^
•f- ro
S- C
cu re
>
0 J=
(J O
CU ro
o£ eu



•
re
































*
o
o
	 1
UJ
s-
o
(4_

cu
c
re
a.
0
s_
o
S-
o
J^
(J
1
1 — C
1
O
E
o
s_
CQ .
i -a
CM C
o
r, CX •
O oo T3
i— i a> cu
a. s- c
• ^
i- 4-> E
O O S-
<+- c cu
4^
CU OO CU
e cu -a
cu o
N T3 4->
c o
cu s- c
JO 0
0 4-> II
S- 0'
o a> •
3 4-> O
r— eu •
i ' f*"i "^



• •
ja o
502.2-27

-------
TABLE 3.  RETENTION TIMES FOR VOLATILE ORGANIC COMPOUNDS ON
           PHOTOIONIZATION DETECTOR (PID)  AND  ELECTROLYTIC
           CONDUCTIVITY DETECTOR(ELCD)  FOR COLUMN  2
                             PID
ECLD

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
AnalvteD
Di chl orodi f 1 uoromethane
Chloromethane
Vinyl Chloride
BromomethanE
Chloroethane
Tri chl orof 1 uoromethane
1,1-Dichloroethene
Methyl ene Chloride
trans-1 , 2-Di chl oroethene
1,1-Di chloroethane
2,2-Dichloropropane
cis-1, 2-Di chl oroethene
Chloroform
Bromochl oromethane
1,1, 1-Tri chl oroethane
1, 1-Dichloropropene
Carbon Tetrachloride
1, 2-Di chl oroethane
Benzene
Trichloroethene
1 , 2-Di chl oropropane
Bromodi chl oromethane
Dibromomethane
Cis-l,3-Dichloropropene
Toluene
Trans-1, 3-Di chl oropropene
1 , 1 , 2-Tri chl oroethane
1 , 3-Di chl oropropane
Tetrachl oroethene
Di bromochl oromethane
1,2-Dibromoethane
Chlorobenzene
1,1,1, 2-Tetrachl oroethane
Ethyl benzene
m-Xyl ene
p-Xylene
o-Xylene
Styrene
Isopropyl benzene
Bromoform
1,1,2, 2-Tetrachl oroethane
1 , 2, 3-Tri chl oropropane
n-Propyl benzene
Bromobenzene
RTfmin}8
-(c)

8.57
_
_
-
14.46
-
17.61
-
-
21.52
-
-
-
24.07
-
-
25.06
2 7; 99
-
-
-
30.40
31.58
32.11
-
-
33.85
-
_
36.76
-
36.92
37.19
37.19
38.77
38.90
40.04
-
-
-
41.51
41.73
RSD


0.06



0.08

0.02


0.02



0.01


0.01
0.01



0.01
0.01
0.01


0.01


0.01

0.01
0.01
0.01
0.01
0.01
0.01


0.01
0.01
0.01
RTYmin}3
7.36
8.09
8.58
10.39
10.74
11.85
14.47
16.46
17.62
19.25
21.36
21.52
22.08
22.69
23.53
24.08
24.47
24.95

27.15
27.73
28.57
28.79
30.41

32.13
32.69
33.57
33.86
34.58
35.29
36.87
36.87

_
_
_
—
_
40.19
40.64
41.18

41.75
RSD
0.06
0.06
0.08
0.06
0.05
0.07
0.07
0.04
0.03
0.03
0.03
0.02
0.02
0.02
0.02
0.02
0.02
0.01

0.01
0.01
0.02
0.01
0.02

0.01
0.01
0.01
.0.01
0.01
0.01
0.01
0.01






0.01
0.01
0.01

0.01
                         502.2-28

-------
                             TABLE 3  (CONTINUED)
                                      PID
        Anal vie*
 RKminr  RSD
 Internal  Standards
   1-Chloro-2-Fluorobenzene
37.55
0.01
                           ECLD
          RKminr  RSD
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
1 , 3 , 5-Tri methyl benzene
2-Chlorotoluene
4-Chlorotoluene
tert-Butyl benzene
1,2, 4-Trimethyl benzene
sec-Butyl benzene
p-Isopropyl toluene
1 , 3-Di chl orobenzene
1 , 4-Di chl orobenzene
n-Butyl benzene
1,2-Dichlorobenzene
1 , 2-Di bromo-3-Chl oropropane
1,2, 4-Tri chl orobenzene
Hexachl orobutadi ene
Naphthalene
1, 2, 3-Tri chl orobenzene
42.08
42.20
42.36
43.40
43.55
44.19
44.69
45.08
45.48
46.22.
46.88

53.26
53.86
54.45
55.54
0.01
0.01
0.01
0.01
0.01
0.01
0.01
0.01
0.01
0.01
0.01
-
0.01
0.01
0.01
0.01

42.21
42.36
-
-
-
-
45.09
45.48
-
46.89
49.84
53.26
53.87
-
55.54

0.01
0.01




0.01
0.01

0.01
0.01
0.01
0.01

0.01
37.56
0.01
a. Column and analytical  conditions are described  in  Sect.  6.3.4.
b. Number refers to peaks in Figure 502.2-2.
c. - Dash indicates detector does not respond.
                                   502.2-29

-------






Q
O
I
••J
UJ






• C
•o o
CO •£ •—
CO ^?
I — >
cu cu
a: o

>.
cu i- <~
en cu — i
CO > ~-
S- O O
cu o a
> cu •—


CM in 10 o «*•
CO CM CM «3- CM





^-*,
_Q ID LfJ l£> CO r--
- — •'Cn Cn Cn Cn Cn 1
1


                    Q O
                    •Z. =3
              a
              t—i
              a.
                 •a  o
                 4-> •!—
                 CO -4->
                     to
                   ••r-

                  cu  cu
                 Q; O
                 en cu
                 co  >
                 S-  O D
                 cu  a
LU
                              ?—I  O r-H O i—i           O O •—lOi—I O O CM O i—i i—lOOOCMOOOOOOOO
                            i	iii	
                              ooooo           ozooooooooooooooooooooz::
                                                   I   I
                                                         CMCMCO«d-CMCMCMCMCMCNjrOCMCMCMCOCMi—(
                                                                             cr* cr*
                                                                                                       cr>
                                                                                                              00
                                                                                                                    IO
                                                                                                                        COOOOID
                                                                                                                        cr» csS o cr>
o o

o o
IO •—i
                                   I   I    I   I
                                               co ro 10
                                               o o o
                                 CM
                                 o
                                               o o o
                                CO CM
                                o o
                                                                         o o
                                               r-H IO
                                                                                 I    I   I
                                                                  CO CM CO
                                                                  o o o
                                                                                             o o o
                                                                                                        I    I   I
                                                          O CO CO
                                                          r-H O O
                                                                                      000
                                                                                                                            I    I   I
                                                          I   I   I
                                                                      IO LO
                                                                              I   I   I
                                                                                      CM f»- CO
                                                                                                 I    I   I
r^ co
en en
i   i  in to co   i  co  i
     Cn Cn Cn     Cn
                                             I  «=t- I--
                                               cr> en
                                                      i    i   i   i  f-- r-. f-.   i
                                                                                    i  en
                                                                                             en  i   i
                                      cu
                                   I   C
                                  cu  co
                                  £= J=
                                  fO +->
         cu
         c
      cu cu
      C N
   cu  cu c
      Nl CU .
      c .a i
      cu i—
                                            c:
                                            CO
       CU  O
    CU S  S-
    c o  o
    CU S- i—
    N O J=  E
    C i—  O  S- -M J3
 CU  CU .C -r-  O CU i—
 cja UTS'*- e  >>=3
 cuooooo4->ca
 M  E S  E  e E  3   i
 SZOOOOOCQO
 CU  S- S-  S-  S- S-  I   CU
GO CO PO CO fy^ rt*^  c^  c/)
  cu
 •a
 -r-
i  S-
:  o
11—
I j=
:  o

I  S-
• 4->
^ CU
                                       cu

                                       CO
                                       Q.
                                       O

                                       CL
                                       O  CU
                                       S-  C
                                       o   +J
                                                                   ccccucu
                                                                   CUCUCUEC
                                                                   NNNOaJ
                                                                                       cu
                                                                                       O
                                                                                   i— O
                                                                                           cu
                                                             CU+JOCUOOS-U
                                                      CO
                                                       I
                                  o o
                                  S- S-
                                  o o
                                                                c_>
                                 o  o -I—  E
                                i— n- Q  O
                                J= J=  I   S-
                                C_3 C_5 CM XI
                                 i
                                CM
    C JD XI XI =3  CU
    cO O  O  O r—  O
    r: s-  i-  s- 1-  i-
    4-> O  O  O -I-  O
    CU r— r— i— ~O r-~
xi  E x: jc j= ox:
•i-  O O  O  O S-  O
Q  E -r- -r- -r- O ••-
 I   O Q Q Q i— Q
CM  S-  i   i    i  JT  i
  "XI CM CO «* O i—I
           cu
           c
       cu  cu
       C -C
       CU 4->
       .c  cu  cu cu cu
 CU  CU -4->  O  C £= C
 c  c: cu  s-  co co co
 ce  cu o  o  o. a. a.
JC ^= S- r—  O O O

 cu  cu i—  o  a. a. a.
 o  o x: -r-  o o o
 S-  S- O O  S- S- S-
 O  O -r-  I   O O O
i— r— O CM r— p— r—
                                                                                                               O  O CM i-H  O  O U
                                                                                                                         I
                                                                                r—I Q    Q
                                                                                                      (Qi
                                                                                                              QQr-i
                                                                                                               I    I   I   C  I    I   I
                                                                                                              CM •— I  in  COCMCOCM


                                                                                                              i-Hi— I  r-1r-(CM
                                                              502.2-30

-------



^
•**.
	 1 c
Q =
Z *"
• C
T3 O

OO •+-> --
rB &S
i — >
cu cu
O CC 0
CJ>
	 1
4 , i «^^
\ 1 1 ^>)
 ^*.
i~ o a
^, cu o a
/—•» > CU ' —
° £
CU CU
o: Q
Q

CL >>
CU i. —
CD CU —1
rB > ^-
S- O D
cu o a
> cu •*-












r

cu "
•4-^ •
> t
"~a r
 C

^ ** 1 t^
3 t— 1 01 1—
-I Jl § .C1
-To t^ LU :




°.| |°.| QOO 00<
o o 2:00 ooc



f*^1"1 CMCOOI co LO u
f'J CO CM CM r— I CM CM C





o co o f-- oo 10 u
I*** 1 1 CD 1 1 1 CTi CD CT» 1 £r^ fT^ r
cn | ^_.



—-

CJl CM CM CMOOO -^-CMLOIO
OOO O O r— i OOOO
ooo ooo oooo





CVJ «— « CNJ CM •— < i— 1 ,_Hr_,Ojo^






LO t11^^ tO CD f*^ to J1*-^ rrt 10 .-^
O^ O^ O^ I CT\ G*i O*i 1 1 fT» £T\ rr\ i^T»
i wi vj> Ui O» I









CU O)
(& ffl
^= ^: cu ' OOO
= =4->jC: Nl S-S-0) r-^r-1
OCUi — OCUC +J+JQ _cr _gr r-
o ^^ >> ecu cucu&- ocjo
o >^ Q ^ ^_ ^^ * , • . ^- *r~ *^ '^~
~~ <^ SH CU ffl ^«\ flj C\J C\J '. rtl , ,** ,^"
— Q ^^ (.• ^^ :II ^^^ "^ * ' ' *~
O S- O >»-!-> O CU i— i CM ro CUOO-^-t— i
r8O-01.CJi:S-S- --S-u «««
c "— ' Q.S: ~z. c oo ^H ^-t i — i — •_T_r_J>




D O O CO O t-H
D O O O O O ' ' '



D IO CM O O to
vJ i— 1 i— 1 ID CM CM






\ 2? 5? ^r ^^ ^^
Oi Ol C3l CD 1 1 Cn III1
fr~*




CO CM CO •— « CM CM CM
o o o o o o o
o o o o o o o




r-- o to ~H ,-H -H en








r^* to 00 LO OO 00 00
i en i i en en en en en en











cu cu cu eu
cu = c: c c
sz nj  CU i- CU CU
CU E O..Q J3
O CU O O r— i —
5- C S- S- >> >> CU
o eu o o _e .c -o
i — -£T 3 i — +J ^J -r-
0 CU £1 "o E E O

t^* * *- *- ^ i- .c cu cu cu
^~ O O p~ ^— ^— O C C C
pjj^^''' CUQJCU
-_(J 0 - - ~'>>'>^'^)
^ II T ^ ^ ^.^ ^ ^f ^
a.
0
1— 1
 -o
cu o
00 5
i g
0 E
S-
t*- T3
s-
TD (O
CU "O
C C
•I- 
0) X
2 CU
V) >>
c .a
o
•r- T3
•M CU
<« c

> E
cu s-
-o cu
.4^
-o cu
S- TD
ns
•a cu
c s-
rfl CU
•M S .
n -a
00 £=
cu cu o
> -i- 0- •
r- i- 00 "O
+J CU CU CU
ro > i. c
— O •!-
CU O +J E
1- CU 0 S-
0£ C CU
O *4-^
C • 00 CU
fO CO CU ^3
-t-> o
CU i— O
"- "> S- C
s- = o
CU nj ^J u
> u
o -r: a>
U O +J Q
0) ret CU •
ai cu Q 2:

. .
                            o e o.
502.2-31

-------
   OPTIONAL
   FOAM
   TRAP
KIN.  _
0. 0. EXIT
X IN.
                    14MM 0. 0.
                  INLET KIN.
                       o.o.
  10MM GUSS F»T
  MEDIUM POROSITY
                            INLfT

                           SYRJNGH VALVE
                      17C«. M GAUGE SYWNGS
  . 0. 0. RUBBER SEPTUM

 ~1ftJM. 0. 0.   1/16 IN. O.D.
                             IN. 0. 0.
                                               STER
                                      13* MOLECULAR
                                      SIEVE PURGE
                                      GAS FILTER
                                        PURGE GAS
                                        ROW
                                        CONTROL
            FIGURE 1.  PURGING DEVICE
                    502.2-32

-------
     PACKING PROCEDURE
            CONSTRUCTION
    GLASS
    WOOL
ACTIVATE)
     7A/FOOT
   RESISTANCE
 •IRE WRAPPED
     SCUD
(DOUBLE LAYER)
                     7^/FOOT
                   RESISTANCE
                  WIRE WRAPPED
                        SOLID
                 (SINGLE LAYER)
                 INLET
COMPRESSION
FITTING NUT
AND FERRULES


 THERMOCOUPLE/
 CONTROLLER
 SBtSOft
                          aECTRONIC
                          TBIPBATUREj
                          CONTROL
                          AND
                          PYROMETER
                       TUBING 2SCM
                       0.105 IN. I.D.
                       0.12S IN. 0,0,
                       STAINLESS STER
     FIGURE  2.  TRAP PACKINGS AND CONSTRUCTION TO INCLUDE
               DESORB CAPABILITY
                    502.2-33.

-------
 PS
 s
8
g
*^
1

-------
o_

-------
                                                                   1
THIS PAGE LEFT BLANK INTENTIONALLY
            502.2-36

-------
METHOD 504.1   1,2-DIBROMOETHANE (EDB), l,2-DIBROMO-3-CHLORO-
               PROPANE (DBCP), AND 1,2,3-TRICHLOROPROPANE (123TCP) IN
               WATER BY MICROEXTRACTION AND GAS CHROMATOGRAPHY
                                 Revision 1.1
                          Edited by J.W.  Munch (1995)
            T. W. Winfield - Method 504, Revision  1.0  (1986)

            T. W. Winfield - Method 504, Revision  2.0  (1989)

            James W.  Eichelberger - Method  504.1,  Revision  1.0  (1993)
                     NATIONAL EXPOSURE RESEARCH LABORATORY
                       OFFICE OF RESEARCH  AND  DEVELOPMENT
                      U.S.  ENVIRONMENTAL PROTECTION  AGENCY
                             CINCINNATI, OHIO  45268
                                    504.1-1

-------
                                  METHOD 504.1

        1,2-DIBROMOETHANE (EDB), l,2-DIBROMO-3-CHLOROPROPANE (DBCP), AND
                   1,2,3-TRICHLOROPROPANE  (123TCP)  IN WATER  BY
                    MICROEXTRACTION AND GAS CHROMATOGRAPHY
      SCOPE AND APPLICATION
      1.1
 This method (1-3) is applicable to the determination of the
 following compounds in finished drinking water and groundwater:
           Analvte

           1,2-Dibromoethane
           1,2-Di bromo-3-Chloropropane
           1,2,3-Trichloropropane
                              Chemical Abstract Services
                                    Registry Number

                                       106-93-4
                                        96-12-8
                                        96-18-4
     1.2
     1.3
 For compounds other than the above mentioned analytes,  or for other
 sample sources,  the analyst must demonstrate the usefulness of the
 method by collecting precision and accuracy data on actual  samples
 and provide qualitative confirmation of results by gas
 chromatography/mass spectrometry (GC/MS)  (4).

 The experimentally determined method detection limits (MDL) (5)  for
 EDB and DBCP were calculated to be 0.01 0g/L and the MDL for 123TCR
 was calculated to be 0.02 /tg/L.  The method has been useful for
 these  anal'/tes over a concentration range  from approximately 0.03  to
 200 /jg/L.   Actual  detection limits are  highly  dependent  upon the
 characteristics  of the gas chromatographic system used.
2.   SUMMARY OF METHOD
     2.1
     2.2
     2.3
Thirty-five  mL  of  sample  are  extracted with  2  mL  of  hexane.  Two /zL
of the  extract  are then injected  into a  gas  chromatograph equipped
with  a  linearized  electron  capture detector  for separation and
detection.   Analytes  are  quantitated using procedural  standard
calibration  (Sect.  3.12).

The extraction  and  analysis time  is 30 to 50 min  per sample
depending upon  the  analytical conditions chosen.

Confirmatory evidence should  be obtained for all  positive results.
This data may be obtained by  using retention data from a dissimilar
column, or when concentrations are sufficiently high by GC/MS.
Purge and trap techniques using Methods 502.2 or  524.2 may also be
used.  Confirmation of all positive results of EDB are especially
important, because  of the potential for misidentification of
dibromochloromethane  (DBCM) as EDB.
                                   504.1-2

-------
3.   DEFINITIONS

     3.1  LABORATORY DUPLICATES (LD1 and LD2) — Two aliquots of the same
          sample taken in the laboratory and analyzed separately with
          identical procedures.  Analyses of LD1 and LD2 indicate the
          precision associated with laboratory procedures, but not with sample
          collection, preservation, or storage procedures.

     3.2  FIELD DUPLICATES (FD1 and FD2) — Two separate samples collected at
          the same time and place under identical circumstances and treated
          exactly the same throughout field and laboratory procedures.
          Analyses of FD1 and FD2 give a measure of the precision associated
          with sample collection, preservation and storage, as well as with
          laboratory procedures.

     3.3  LABORATORY REAGENT BLANK  (LRB) — An aliquot of reagent water or
          other blank matrix that is treated exactly as a sample including
          exposure to all glassware, equipment, solvents, reagents, internal
          standards, and surrogates that are used with other samples.  The LRB
          is used to determine  if method analytes or other interferences are
          present in the laboratory environment, the reagents, or the
          apparatus.

     3.4  FIELD REAGENT BLANK  (FRB) — An aliquot of reagent water or other
          blank matrix that is  placed in a sample container in the laboratory
          and treated as a sample in .all respects, including shipment to the
          sampling site, exposure to sampling site conditions, storage,
          preservation and all  analytical procedures.  The purpose of the FRB
          is to determine if method analytes or other interferences are
          present  in the field  environment.

     3.5  INSTRUMENT PERFORMANCE CHECK SOLUTION  (IPC) — A solution of one or
          more ,method analytes, surrogates,  internal standards, or other test
          substances used to evaluate the performance of the instrument system
          with respect to a defined set of criteria.

     3.6  LABORATORY FORTIFIED  BLANK  (LFB) — An aliquot of reagent water or
          other blank matrix to which known  quantities of the method  analytes
          are added  in the laboratory.  The  LFB  is analyzed exactly like a
          sample,  and its purpose  is to determine whether the methodology is
          in control, and whether the laboratory is capable of making accurate
          and precise measurements.

     3.7  LABORATORY FORTIFIED  SAMPLE MATRIX (LFM) — An  aliquot of an
          environmental  sample  to which known quantities of the method
          analytes  are added  in the laboratory.  The LFM  is analyzed  exactly
          like a  sample, and  its purpose  is  to determine whether the  sample
          matrix  contributes  bias  to the  analytical results.  The  background
          concentrations of the analytes  in  the  sample matrix must be
          determined in  a separate  aliquot  and the measured values in the LFM
          corrected  for  background  concentrations.


                                    504.1-3

-------
      3.8  STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing
           one or method analytes prepared in the laboratory using assayed
           reference materials or purchased from a reputable commercial  source.

      3.9  PRIMARY DILUTION STANDARD SOLUTION (PDS)  - A solution of several
           analytes prepared in the laboratory from  stock standard solutions
           and diluted as needed to prepare calibration solutions and other
           needed analyte solutions.

      3.10 CALIBRATION STANDARD (CAL)  - A solution  prepared from the primary
           dilution standard solution  and stock standard solutions of the
           internal standards and surrogate analytes.   The CAL  solutions are
           used to calibrate the instrument response with respect to analyte
           concentration.                                                J

      3.11  QUALITY CONTROL  SAMPLE (QCS)  -  A  solution  of method  analytes of
           known  concentrations that is  used  to  fortify an aliquot of LRB or
           sample matrix.   The  QCS  is  obtained  from  a  source  external  to the
           laboratory  and different from the  source  of calibration standards
           It  is  used  to check  laboratory performance  with externally prepared
           test materials.                                             K  M«ICU

      3.12  PROCEDURAL  STANDARD  CALIBRATION  -  A calibration method where
           aqueous  calibration  standards  are  prepared  and  processed  (e g
           purged  extracted,  and/or derivatized) in  exactly the  same  manner as
           a sample.   All steps  in the process from  addition of  sampling
           preservatives through  instrumental analyses  are  included  in the
           calibration.  Using  procedural standard calibration compensates for
           any  inefficiencies in the processing procedure.

4.   INTERFERENCES

     4.1   Impurities contained in the extracting solvent usually account for
          the majority of the analytical problems.  Solvent blanks should  be
          analyzed on each new bottle  of solvent before use.  Indirect daily
          checks on the extracting solvent are obtained by monitoring the
          reagent water blanks (Sect.  7.2.4).  Whenever an interference  is
          noted in the reagent water blank, the analyst should  reanalyze the
          extracting solvent.  Low level interferences generally can be
          removed by distillation or  column chromatography (3).   When a
          solvent is purified,  preservatives  put into  the solvent by the
          manufacturer are  removed  thus  potentially  making the  shelf-life
          short.   It is generally more economical  to obtain a new source of
          solvent.  Interference-free  solvent is defined as a solvent
          containing less than  the  MDL of an  individual analyte  interference
          Protect interference-free solvents  by  storing in an area free  of
          organochlorine  solvents.

    4.2  This  liquid/liquid extraction  technique  efficiently extracts a wide
         boiling range of  non-polar organic  compounds and,  in addition,
         extracts polar organic  components of the sample  with varying
         efficiencies.                                                  .

                                   504.1-4

-------
     4.3  Dibromochloromethane is a common disinfection byproduct in
          chlorinated drinking waters that frequently occurs at relatively
          high concentrations.  DBCM can elute very close to EDB, and a high
          concentration of DBCM may mask a low concentration of EDB, or be
          misidentified as EDB.  Therefore, special care should be taken in
          the identification and confirmation of EDB.

     4.4  It is important that samples and working standards be contained in
          the same solvent.  The solvent for working standards must be the
          same as the final solvent used in sample preparation.  If this is
          not the case, chromatographic comparability of standards to sample
          may be affected.

5.   SAFETY

     5.1  The toxicity and carcinogenicity of chemicals used in this method
          have not been precisely defined; each chemical should be treated as
          a potential health hazard, and exposure to these chemicals should be
          minimized.  Each laboratory is responsible for maintaining awareness
          of OSHA regulations regarding safe handling of chemicals used in
          this method.  Additional  references to laboratory safety are
          available (5-7) for the information of the analyst.

     5.2  EDB, DBCP, and 123TCP have all been tentatively classified as known
          or suspected human or mammalian carcinogens.   Pure standard
          materials and stock standard solutions of these compounds should be
          handled in a hood or glovebox.  A NIOSH/MESA approved toxic gas
          respirator should be worn when the analyst handles high
          concentrations of these toxic compounds.

6.   EQUIPMENT AND SUPPLIES  (All specifications are suggested.  Catalog
     numbers are included for illustration only.)

     6.1  SAMPLE CONTAINERS — 40-mL screw cap vials each equipped with a
          Teflon-lined cap.  Individual vials shown to contain at least 40.0
          mL can be calibrated at the 35.0 mL mark so that volumetric, rather
          than gravimetric, measurements of sample volumes can be performed.
          Prior to use, wash vials and septa with detergent and rinse with tap
          and distilled water.  Allow the vials and septa to air dry at room
          temperature, place in a 105°C oven for one hr, then remove and allow
          to cool in an area free of organic solvent vapors.

     6.2  VIALS, auto sampler, screw cap with Teflon faced septa, 1.8 mL.

     6.3  MICRO SYRINGES — 10, 25, and 100 /tL.

     6.4  PIPETTES -- 2.0 and 5.0 mL transfer.

     6.5  STANDARD SOLUTION STORAGE CONTAINERS — 15-mL bottles with Teflon
          lined screw caps.

     6.6  GAS CHROMATOGRAPHY SYSTEM

                                    504.1-5

-------
          6.6.1  The gas chromatograph must be capable of temperature
                 programming and  should be equipped with a linearized electron
                 capture detector and a capillary column split/split!ess
                 injector.

          6.6.2  Two gas chromatography columns are recommended.  Column A
                 provides separation of the method analytes without
                 interferences from trihalomethanes (Sect. 4.3).  Column A
                 should be used as the primary analytical column unless
                 routinely occurring analytes are not adequately resolved.
                 Column B is recommended for use as a confirmatory column when
                 GC/MS confirmation is not viable.  Retention times for the
                 method analytes  on these columns are presented in Table 1.

          6.6.3  Column A (primary column) — DB-1, 30 m x 0.25 mm ID, 1.0 /im
                 film thickness fused silica capillary column or equivalent.
                 The linear velocity of the helium carrier gas should be about
                 25 cm/sec at 100°C.  The column temperature is programmed to
                 hold at 40°C for 4 min, to increase to 240°C at 10°C/min, and
                 hold at 240°C for 5 min or until all expected compounds have
                 eluted.

          6.6.4  Column B (alternative column) — DB-624, 30 m x 0.32 mm ID,
                 1.8 (im film thickness fused silica capillary column or
                 equivalent.  The linear  velocity of the helium carrier gas
                 should be about  25 cm/sec at 100°C.  The column temperature
                 is programmed as described in Sect.6.6.3.

7.   REAGENTS AND STANDARDS

     7.1  REAGENTS

          7.1.1  Hexane extraction solvent — UV Grade, distilled in glass

          7.1.2  Methyl alcoliol — ACS Reagent Grade, demonstrated to be free
                 of method analytes above the MDLs.

          7.1.3  Sodium chloride, NaCl — ACS Reagent Grade - For pretreatment
                 before use, pulverize a batch of NaCl and place in a muffle
                 furnace at room  temperature.  Increase the temperature to
                 400°C for 30 min.  Place in a bottle and cap.

          7.1.4  Sodium thiosulfate, Na2S20,, ACS Reagent Grade — For
                 preparation of solution (40 mg/mL), dissolve 1 g of Na2S203
                 in reagent water and bring to 25-mL volume in a volumetric
                 flask.

     7.2  REAGENT WATER — Reagent water is defined as water free of
          interferences above the analyte MDLs.

          7.2.1  Reagent water can be generated by passing tap water through a
                 filter bed containing activated carbon.  Change the activated

                                    504.1-6

-------
            carbon when there is evidence that volatile organic compounds
            are breaking through the carbon.

     7.2.2  A Mi Hi pore Super-Q Water System or its equivalent may be
            used to generate deionized reagent water.

     7.2.3  Reagent water may also be prepared by boiling water for
            15 min.  Subsequently, while maintaining the temperature at
            90°C, bubble a contaminant-free inert gas through the water
            at 100 mL/min for 1 hr.  While still hot, transfer the water
            to a narrow mouth screw cap bottle with a Teflon seal.

     7.2.4  Test reagent water each day it is used by analyzing it
            according to Sect. 11.

7.3  STOCK STANDARD SOLUTIONS — These solutions may be purchased as
     certified solutions or prepared from pure standard materials using
     the following procedures:

     7.3.1  Place about 9.8 mL of methanol into a 10-mL ground-glass
            stoppered volumetric .flask.  Allow the flask to stand,
            unstoppered, for about 10 min and weigh to the nearest 0.1
            mg.

     7.3.2  Use a 100-#L syringe and immediately add two or more drops of
            standard material to the flask.  Be sure that the standard
            material falls directly into the methanol without contacting
            the neck of the flask.

     7.3.3  Reweigh, dilute to volume, stopper, then mix by inverting the
            flask several times.  Calculate the concentration in
            micrograms per microliter from the net gain in weight.

     7.3.4  Store stock standard solutions in 15-mL bottles equipped with
            Teflon lined screw caps.  Methanol solutions prepared from
            liquid analytes are stable for at least four weeks when
            stored at 4°C.

7.4  PRIMARY DILUTION STANDARD SOLUTIONS — Use stock standard solutions
     to prepare primary dilution standard solutions that contain all
     three analytes in methanol.  The primary dilution standards should
     be prepared at concentrations that can be easily diluted to prepare
     aqueous calibration standards (Sect. 10.1.1) that will bracket the
     working concentration range.  Store the primary dilution standard
     solutions with minimal headspace and check frequently for signs of
     deterioration or evaporation, especially just before preparing
     calibration standards.  The storage time described for stock
     standard solutions in Sect. 7.3.4 also applies to primary dilution
     standard solutions.
                               504.1-7

-------
     7.5  LABORATORY FORTIFIED BLANK  (LFB) SAMPLE CONCENTRATE (0.25 fig/ml) —
          Prepare an LFB sample concentrate of 0.25 /ig/mL of each analyte from
          the stock standard solutions prepared in Sect. 7.3.

     7.6  MDL CHECK SAMPLE CONCENTRATE (0.02 /ig/mL) — Dilute 2 mL of LFB
          sample concentrate (Sect. 7.5) to 25 mL with methanol.

8.   SAMPLE COLLECTION. PRESERVATION. AND STORAGE

     8.1  SAMPLE COLLECTION

          8.1.1  Replicate field reagent blanks (FRB) must be handled along
                 with each sample set, which is composed of the samples
                 collected from the same general  sampling site at
                 approximately the same time.  At the laboratory, fill a
                 minimum of two sample bottles with reagent water, seal, and
                 ship to the sampling site along with sample bottles.
                 Wherever a set of samples is shipped and stored, it must be
                 accompanied by the FRB.

          8.1.2  Collect all  samples in 40-mL bottles into which  3 mg of
                 sodium thiosulfate crystals have been added to  the empty
                 bottles just prior to shipping to the sampling  site.
                 Alternatively, 75 /*L of freshly prepared sodium  thiosulfate
                 solution (40 mg/mL) may be added to empty 40-mL  bottles just
                 prior to sample collection.   This dechlorinati'ng agent must
                 be added to each sample to avoid the possibility of reactions
                 that may occur between residual  chlorine and indeterminant
                 contaminants present in  some solvents,  yielding  compounds
                 that may subsequently interfere  with the analysis.  .The
                 presence of sodium thiosulfate will  arrest further formation
                 of DBCM (See Sect.  4.3).

          8.1.3  When sampling  from a water tap,  open the tap and allow the
                 system to flush until the water  temperature has  stabilized
                 (usually about 10 min).   Adjust  the flow to about 500 mL/min
                 and collect  samples from  the flowing stream.

          8.1.4  When sampling  from a well,  fill  a wide-mouth bottle  or beaker
                 with sample,  and carefully fill  40-mL sample bottles.

     8.2  SAMPLE PRESERVATION

          8.2.1   The samples  must be chilled  to 4°C  or less  at the time of
                 collection  and maintained at that temperature until  analysis.
                 Field  samples  that  will not  be received  at  the laboratory on
                 the day of  collection must  be  packaged  for  shipment  with
                 sufficient  ice to ensure  that  they  will  be  <4°C  on arrival  at
                 the laboratory.
                                   504.1-8

-------
     8.3  SAMPLE STORAGE

          8.3.1  Store samples and field reagent blanks together at 4°C until
                 analysis.  The sample storage area must be free of organic
                 solvent vapors.

          8.3.2  Because 1,2,3-trichloropropane has been added to the analyte
                 list in this method and has been found to have a 14 day
                 maximum holding time in studies conducted for Method 524.2
                 (4), all samples must be extracted within 14 days of
                 collection.  Samples not analyzed within this period must be
                 discarded and replaced.  Because of the potential for solvent
                 evaporation, it is preferred that extracts be analyzed
                 immediately following preparation.  When necessary, extracts
                 may be stored in tightly capped vials (Sect. 6.2) at 4°6 or
                 less for up to 24 hr.

9.   QUALITY CONTROL   .

     9.1  Each laboratory that uses this method is required to operate a
          formal quality control program.  The minimum requirements of this
          program consist of an initial  demonstration of laboratory capability
          and an ongoing analysis of field reagent blanks (FRB), laboratory
          reagent blanks (LRB), laboratory fortified blanks (LFB), laboratory
          fortified sample matrix (LFM), and quality control  samples (QCS) to
          evaluate and document data quality.  Ongoing data quality checks are
          compared with established performance criteria to determine if the
          results of analyses meet the performance characteristics of the
          method.

          9.1.1  The analyst must make an initial determination of the method
                 detection limits and demonstrate the ability to generate
                 acceptable precision with this method.  This is established
                 as described in Sect. 9.2.

          9.1.2  In recognition of advances that are occurring in chromato-
                 graphy, the analyst is  permitted certain options to improve
                 the separations or lower the cost of measurements.  Each time
                 such a modification is  made to the method, the analyst is
                 required to repeat the  procedure in Sect. 9.2.

          9.1.3  Each day, the analyst must analyze a laboratory reagent blank
                 (LRB) and a field reagent blank, if applicable (Sect. 8.1.1),
                 to demonstrate that interferences from the analytical system
                 are under control before any samples are analyzed. In .
                 general, background interferences co-eluting with method
                 analytes should be below the method detection limits.

          9.1.4  The laboratory must, on an ongoing basis, demonstrate through
                 the analyses of laboratory fortified blanks  (LFB) that the
                 operation of the measurement system is in control.  This


                                    504.1-9

-------
             procedure is described in Sect. 9.3.  The frequency of the
             LFB analyses is equivalent to 10% of all samples analyzed.

      9.1.5  The laboratory should demonstrate the ability to analyze low
             level samples weekly.  The procedure for low level  LFB
             samples is described in Sect. 9.4.

 9.2  Initial Demonstration of Capability

      9.2.1  Prepare four to seven LFBs at a concentration equal to 10
             times the MDL or at a concentration in the middle of the
             calibration range established in Sect. 10.

      9.2.2  Analyze the LFBs according to the method beginning  in Sect.
             JL Ji •

      9.2.3  Calculate the mean concentration found (X) in /zg/L, and the
             standard deviation of the concentrations in /jg/L,  for each
             analyte.

      9.2.4  For  each analyte,  X should be between 70% and 130%  of the
             true value.   The RSD should be 20% or less.  If the  results
             for  all  three analytes  meet these criteria,  the system
             performance  is  acceptable.   If any analyte fails to meet the
             criteria,  correct  the source of the problem and repeat the
             test.

             CAUTION:  No  attempts to  establish low detection limits
             should  be  made  before instrument optimization and adequate
             conditioning  of both the  column  and the GC system.
             Conditioning  includes the processing of LFB  and LFM samples
             containing moderate analyte concentrations.

      9.2.5   Determination of MDL.   Prepare 4-7 LFBs at a low
             concentration.   Use the concentrations  in  Tables 2  and  3  as a
             guideline, or use  calibration  data obtained  (Sect.  10)  to
             estimate a concentration  for  each  analyte  that  will  produce a
             chromatographic peak with a  3-5  signal  to  noise ratio.   It is
             recommended that LFBs for determination  of the  MDL  be
             prepared and  analyzed over  a  period  of  several  days,  so  that
             day  to day variations will  be  reflected  in the  precision
             data.

     9.2.6   Analyze  the LFBs as  directed  in  Sect.  11.  Calculate  the mean
             amount recovered and  the  standard  deviation  of  these
             measurements.   Use  the standard  deviation  and the equation in
             Sect. 13 to calculate the MDL.

9.3  Assessing Laboratory Performance.  The  laboratory must  demonstrate
     that the measurement system  is in control by analyzing  an  LFBs of
     the analytes at 0.25 fj.g/1  concentration level. This must be
                              504.1-10

-------
     demonstrated on a frequency equivalent to 10% of the sample load, or
     1 per batch of samples extracted, whichever is greater.

     9.3.1  Prepare an LFB sample (0.25 jtig/L) by adding 35 jil of LFB
            concentrate (Sect. 7.5) to 35 mL of reagent water in a 40-mL
            bottle.

     9.3.2  Immediately analyze the LFB sample according to Sect. 11 and
            calculate the recovery for each analyte.  The recovery should
            be between 70% and 130% of the expected value.

     9.3.3  If the recovery for either analyte falls outside the
            designated range, the analyte fails the acceptance criteria.
            A second LFB containing each analyte that failed must be
            analyzed. .Repeated failure, however, will confirm a general
            problem with the measurement system.  If this occurs, locate
            and correct the source of the problem and repeat the test.

     9.3.4  Since this LFB is prepared in the same manner as a
            calibration verification standard, this LFB data can also be
            used to satisfy the calibration requirement in Sect. 10.1.4.

9.4  Assessing Laboratory Sensitivity.  The laboratory should demonstrate
     the ability to analyze low level samples for EDB and DBCP weekly.

     9.4.1  Prepare an MDL check sample (a LFB fortified at 0.02 /jg/L)
            and immediately analyze according to the method in Sect. 11.

     9.4.2  The instrument response must indicate that the laboratory's
            MDL is distinguishable from instrument background signal.  If
            it is not, correct the problem (increase sensitivity) and
            repeat Sect. 9.4.1.

     9.4.3  For each analyte, the recovery must be between 60% and 140%
            of the expected value.  These criteria are looser than those
            in Sect. 9.2.4 and 9.3.2 because of the low concentration.

     9.4.4  When either analyte fails the test, the analyst should repeat
            the test for that analyte.  Repeated failure, however, will
            confirm a general problem with the measurement system or
            faulty samples and/or standards.  If this occurs, locate and
            correct the source of the problem and repeat the test.

9.5  Assessing Matrix Effects.  At least once in every 20 samples,
     fortify an aliquot of a randomly selected,routine sample with known
     amounts of the analytes.  The added concentration should not be less
     than the background concentration of the sample selected for
     fortification.  To simplify these checks, it would be convenient to
     use LFM concentrations ~10X MDL.  Over time, recovery should be
     evaluated on fortified samples from all routine sources.  Calculate
     the percent recovery (R,-) for each  analyte,  corrected for background
     concentrations measured in the unfortified sample.  If the recovery

                              504.1-11

-------
          of any such analyte falls outside the range of ± 35% of the
          fortified amount, and the laboratory performance for that analyte is
          shown to be in control (Sect. 9.3), .the recovery problem encountered
          with the dosed sample is judged to be matrix related, not system
          related.  The result for that analyte in the unfortified sample is
          labeled suspect/matrix to inform the data user that the results are
          suspect due to matrix effects.

     9.6  It is highly recommended that a laboratory establish its
          ability to distinguish DBCM from EDB.  This is
          particularly important if samples from chlorinated sources
          or unfamiliar sources are to be analyzed (Sect. 4.3).
          Standards of DBCM should be analyzed periodically to
          establish its retention time relative to that of EDB.
          When evaluating this retention time difference,  the
          analyst should keep in mind that DBCM is likely to be
          present in concentrations much larger than EDB, and that
          the ability to detect EDB may deteriorate with increasing
          DBCM concentration.

     9.7  At least quarterly, a quality control sample (QCS) should be
          analyzed.  If measured analyte concentrations are not of acceptable
          accuracy, check the entire analytical procedure to locate and
          correct the problem source.

     9.8  It is recommended that the laboratory adopt additional  quality
          assurance practices for use with this method.  The specific
          practices that are most productive depend upon the needs of the
          laboratory and the nature of the samples.  Field duplicates may be
          analyzed to assess the precision of the environmental measurements.
          Whenever possible, tf.e laboratory should analyze standard reference
          materials and participate in relevant performance evaluation
          studies.

10.  CALIBRATION AND STANDARDIZATION

     10.1 CALIBRATION AND STANDARDIZATION

          10.1.1 At least three calibration standards are needed;  five are
                 recommended.  Guidance on the number of standards is as
                 follows:  A minimum of three calibration standards are
                 required to calibrate a range of a factor of 20  in
                 concentration.  For a factor of 50 use at least  four
                 standards, and for a factor of 100 at least five  standards.
                 The lowest standard should represent analyte concentrations
                 near,  but above,  their respective MDLs.   The remaining
                 standards should bracket the analyte concentrations expected
                 in the sample extracts, or should define the working range of
                 the detector.

          10.1.2 To prepare a calibration standard (CAL), add an  appropriate
                 volume of a primary dilution standard solution to an aliquot

                                   504.1-12

-------
                 of reagent water in a volumetric flask.  If < 20 /iL of a
                 standard is added to the reagent water, poor precision may
                 result.  Use a 25-/JL micro syringe and rapidly inject the
                 alcoholic standard into the expanded area of the filled
                 volumetric flask.  Remove the needle as quickly as possible
                 after injection.  Mix by inverting the flask several times.
                 Discard the contents contained in the neck of the flask.
                 Aqueous standards should be prepared fresh and extracted
                 immediately after preparation Unless sealed and stored
                 without headspace as described in Sect. 8.  Alternatively,
                 measure a 35-mL volume of reagent water in a 50-mL graduated
                 cylinder and transfer it to a 40-mL sample container (Sect.
                 6.1).  Use a micro syringe to inject the standard into the
                 reagent water.  Cap and mix gently.

          10.1.3 Analyze each calibration standard according to Sect. 11 and
                 record the peak height or area response from each standard.
                 Create a calibration curve by plotting peak area response
                 versus the concentration in the standard.  Alternatively, if
                 the ratio of concentration to response (calibration factor)
                 is a constant over the working range (< 20% relative standard
                 deviation), linearity through the origin can be assumed and
                 the average ratio or calibration factor can be used in place
                 of a calibration curve.

          10.1.4 Verify the calibration daily by the analysis of 1 or more
                 calibration standards for each 12 hr shift of operation.  It
                 is recommended that a calibration standard be analyzed at the
                 beginning of each period of operation, and also at the end of
                 each period of continuous instrument operation.  Vary the
                 concentration of the calibration standards used for
                 verification, so that several points in the calibration range
                 are verified.  NOTE: The data presented in Tables 1-3 were
                 obtained on a chromatographic system that was calibrated
                 daily.  Because of the sensitivity required in this method it
                 may be necessary for some laboratories to calibrate daily, in
                 order to meet the QC criteria in Section 9.

     10.2 INSTRUMENT PERFORMANCE — Check the performance of the entire
          analytical system daily using data gathered from analyses of
          laboratory reagent blanks and standards.

          10.2.1 Significant peak tailing of the target compounds in the
                 chromatogram must be corrected.  Tailing problems are
                 generally traceable to active sites on the GC column,
                 improper column installation, or problems with the operation
                 of the detector.

11.   PROCEDURE

     11.1 SAMPLE PREPARATION
                                   504.1-13

-------
     11.1.1 Remove  samples  and  standards from  storage and allow them to
            reach room temperature.

     11.1.2 For samples  and field reagent blanks, contained in 40-mL
            bottles, remove the container cap.  Discard a 5-mL volume
            using a 5-mL transfer pipette or 10-mL graduated cylinder.
            Replace the  container cap and weigh the container with
            contents to  the nearest 0.1 g and  record this weight for
            subsequent sample volume determination (Sect. 11.3).  NOTE:
            It is important not to use a graduated cylinder or other
            means to transfer the sample to another container prior to
            extraction.  Loss of volatile compounds will occur each time
            the sample is poured or otherwise  transferred.

11.2 MICROEXTRACTION AND ANALYSIS

     11.2.1 Remove the container cap and add 6 g NaCl (Sect. 7.1.3) to
            the sample.

     11.2.2 Recap the sample container and dissolve the NaCl by swirling
            for about 20 sec.

     11.2.3 Remove the cap and add exactly 2.0 mL of hexane using a class
            A, TD, transfer or automatic dispensing pipette.  Recap and
            shake vigorously for 1 min.  Allow the water and hexane
            phases to separate.  (If stored at this stage, keep the
            container upside down.)

     11.2.4 Remove the cap and carefully transfer 0.5 mL of the hexane
            layer into an autoinjector using a disposable glass pipette.

     11.2.5 Transfer the remaining hexane phase, being careful  not to
            include any of the water phase,  into a second autoinjector
            vial.   Reserve this second vial  at 4°C for a reanalysis if
            necessary.

     11.2.6 Transfer the first sample vial  to an autoinjector set up to
            inject 2 ill portions into the gas chromatograph for analysis.
            Alternatively,  2 /iL portions of samples,  blanks and standards
            may be manually injected,  although an autoinjector is
            recommended.

11.3 DETERMINATION OF SAMPLE VOLUME

     11.3.1 For samples and field blanks,  remove the  cap from the sample
            container.

     11.3.2 Discard the remaining sample/hexane mixture.  Shake off the
            remaining few drops using short,  brisk wrist movements.

     11.3.3 Reweigh the empty container with  original  cap and calculate
            the net weight  of sample by difference to the nearest 0.1 g.

                              504.1-14

-------
                   This  net  weight  (grams)  is  equivalent  to  the  volume  of water
                   (in ml) extracted  (Sect.  12.3).
 12.  DATA ANALYSIS AND CALCI1I

      12.1 Identify the method  analytes  in  the  sample  chromatogram  by  comparing
           the retention time of the  suspect peaks  to  retention  times  of  the
           calibration standards and  the  laboratory control  standards  analyzed
           using identical conditions.  The analyst should use caution  in the
           identification of EDB in samples from chlorinated and unknown
           sources that may contain DBCM  (Sect. 4.3).  Confirmation procedures
           in Section 2.3 should be used  to verify  identification of EDB.

      12.2 Use the calibration curve or calibration factor (Sect  10 1 3) to
           directly calculate the unconnected concentration  (C-)  of each
           analyte in the sample (e.g., calibration factor x response).
           Extracts that contain method analytes beyond the calibration ranqe
           established in Sect.  10., must be diluted and reanalyzed.  Do not
           extrapolate beyond the range of instrument calibration   Use the
           multi-point calibration  established  in Sect.10 for all calculations
           Do not use the daily  calibration verification standard to quantitate
           method analytes  in samples.

      12.3  Calculate  the  sample  volume (V )  as equal to the net sample  weight:
           Vs = gross  weight  (Sect.  11.1.!).- bottle tare  (Sect.  11.3.3).

      12.4  Calculate  the  corrected  sample  concentration as:

           Concentration, /ng/L = C,-  x  T^
                                        S

      12.5  Results  should be  reported  with an appropriate  number  of  significant
           ngures.  Experience  indicates  that three significant  figures may be
           used for concentrations above 99 fig/L, two significant figures  for
          concentrations between 1-99 /ig/L, and 1 significant figure for  lower
          VrfUnCcMLiaUlOnS.

13.   METHOD PERFORMANCE

     13.1 Single laboratory accuracy and  precision  data are presented for the
          three method analytes  in reagent water at concentrations of 0 1 ug/l
          T^hinl Vic*'  H f    L1lSiS the data 9enerated using Column A and
          Table 3 lists  data gathered using Column  B.   The method detection
           imits are  presented in Table 1.  The method detection limits (MDL)
          in  the table were calculated using the formula:

                     MDL  =  S t(rM ,_al ha _ o 99)

                 where:

                 t(n-i,i-aiPha = 0.99) = Student's t value for the 99% confidence
                                   level with n-1 degrees of freedom

                 n  = number  of replicates

                 S  = standard deviation  of replicate analyses.

                                  504.1-15

-------
14.  POLLUTION PREVENTION

     14.1 This method utilizes a microextraction procedure that requires the
          use of very small volumes of hexane, thus making this method safe
          for use by the laboratory analyst and harmless to the environment.
          For information concerning pollution prevention that may be
          applicable to laboratory operations, consult."Less is Better:
          Laboratory Chemical Management for Waste Reduction" available from
          the American Chemical Society's Department of Government Relations,
          and Science Policy, 1155 16th Street N.W., Washington, D.C. 20036.

15.  WASTE MANAGEMENT
     15.1
It is the laboratory's responsibility to comply with all federal,
state, and local regulations governing the waste management,
particularly the hazardous waste identification rules and land
disposal restrictions, and to protect the air, water, and land by
minimizing and controlling all releases from fume hoods and bench
operations.  Also, compliance is required with any sewage discharge
permits and regulations.  For further information on waste
management, consult "The Waste Management Manual for Laboratory
Personnel," also available from the American Chemical Society at the
address in Sect. 14.1.
16.  REFERENCES
     1    Glaze, W.W., Lin, C.C.,  "Optimization of Liquid-Liquid Extraction
          Methods for Analysis of  Organics  in Water,"  EPA-600/S4-83-052,
          January 1984.

     2.   Henderson, J.E.,  Peyton,  G.R.  and Glaze, W.H.(1976).  In
          Identification  and Analysis  of Organic  Pollutants  in Water  (L.H.
          Keith ed.), pp.  105-111.  Ann  Arbor Sci. Pub!., Ann Arbor,  Michigan.

     3.   Richard, J.J.,  G.A. Junk, "Liquid Extraction for Rapid Determination
          of Halomethanes  in Water," Journal AWWA, 69, 62, January  1977.

     4.   Munch, J.W.,   "Method  524.2- Measurement of Purgeable Organic
          Compounds  in Water by  Capillary  Column  Gas Chromatography/  Mass
          Spectrometry"  in Methods for the Determination  of  Organic Compounds
          in Drinking Water: Supplement  3   (1995).  USEPA, National Exposure
          Research Laboratory, Cincinnati,  Ohio 45268.

     5.   Glaser, J.A.   D.L. Forest, G.D.  McKee,  S.A. Quave,  and W.L.  Budde,
          "Trace Analyses for Wastewaters," Environmental Science  and
          Technology,  15,  1426  (1981).

     6.   "Carcinogens-Working with Carcinogens," Department of Health,
          Education,  and Welfare,  Public Health Service,  Center for Disease
          Control, National Institute  of Occupational Safety and Health,
          Publication  No.  77-206,  August 1977.

     7.   OSHA Safety  and Health Standards,  (29CFR1910),  Occupational Safety
          and  Health Administration, OSHA 2206.

     8.   Safety  in  Academic Chemistry Laboratories,  American Chemical Society
          Publication,  Committee on Chemical  Safety,  4th  Edition,  1985.
                                    504.1-16

-------
17.   TABLES. DIAGRAMS. FLOWCHARTS. AND VALIDATION DATA
      TABLE 1.  CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS
                 FOR METHOD ANALYTES USING CONDITIONS IN SECTION 6.6.3
Analvte
EDB
123TCP
DBCP
Retention
Column A
9.37
12.00
17.3
Time. Min
Column B
12.47
15.37
15.0
MDL. ua/L

0.01
0.02
0.01
       MDLs  were  calculated  from 8  replicate  samples  fortified  at
       a  concentration  of 0.04  /ig/L of each analyte.
                                  504.1-17

-------
I
o
o
CO
CO
o
UJ
a.

Q
U
O
CM
03
          §
         o
         CM
          ns
         •o
          O)
          s-
          o
          O)
                   c
                   o
                   cu
                   o
                   o
                  o
                   O)

                   o
                   u
                   cu
                  C£.
                           CO
                           o
                           a.
                           o
                           CO

                           CM
         CO

         Q

         LU
                           =«=


                            CU
                            03

                            O
                            a
                            cu
                            a:
          §
          OJ
         T3
          CU
                   Oi
                   C
                   O
 as
 i~

 c
 cu
 u
 c
 o
o

•a
 cu

 cu

 o
 o
 cu
                            CO

                            CM
                            cc

                            Lu
          a>
          •*->
          n]


          £.
         =«=


          CU
                             a>
                            C£
I-H
1 	
I-H
CM
0
LO
f-
i — I
CM
O
CO
CM
O

i — t

CO
en
CD
0
LO
I-H
|—C
O
00
en
o
i-H
CD

i — I
en
o
CM
CM
O
CM
CO
CM
CM
O
O
CM
O

CM

00
i-H
O
o
en
0
o
I-H
CM
i-H
O

CM
O
ID
i-H
CM
0
I-H
CM
CM
O
to
r-.
CM
o

CO

i-H
0
1 — I
o
0
i-H
o

CO
10
en
I-H
CM
O
ID
CO
i-H
CM
O
1 — 1
i-H
CO
o

^J.

oo
1 — 1
i-H
o
I-H
o
0
CM
i-H
O

-
o
ID
i-H
CM
0
ID
OO
i-H
CM
O
10
i-H
CO
0

LO

CO
oo
o
I-H
o
i-H
o
CO
CO
I-H
o

LO
CM
10
CM
O
1 —
0
CO
CM
O
CO
CO
CO
CM
O

ID

CM
CM
i — i
i-H
0
LO
CO
o
0
00
CM
CM
O

(0
LO
CM
CM
O
oo
LO
CM
CM
O
10
CM
i-H
CO
o

r--.

o
en
0
i — i
o
en
CO
i-H
1 — 1
O
o
CO
ez>

-
CO
00
i-H
CM
o
CO
CM
CM
CM
O
CM
r--.
oo
CM
O

c.
CO

LO
o
1 — 1
i-H
o
00
o
I-H
i-H
o
en
10
i — i
0

e
eo
CO
(0
CO
O
o
o
1 —
o
0
o
ID
10
CM
O
°
I-H
C.
LU
Q
Q
1—
OO
1 — 1
o
o
0
o
o
0
CO
en
o
o
CD
i — i
1
C
LU
Q
Q
1—
OO
0
0
o
CM
0
o
o
o
CM
O
CD
O
0
CM
CD
— 1
CD
=i
cu
	 1
o.
oo
0
o
0
I-H
o
o
o
CD
i-H
o
o
CD
o
i-H
o
	 1
CT1
— 1
Q-
oo
                                                                                              en
                                                                                              o
                                                                                              CO
                                                                                                    un
                                                                                                    10
                                                                                 CO

                                                                                 I-H



                                                                                 CM
                                                                                                    in
                                                                                                    CM
                                                                                              CXL
                                                                                              UJ    O
                                                                                              >    00
                                                                                              o    o;
                                                                                              o
                                                                                              LU    «
                                                                                               LO
                                                                                                     00
                                                                                                     CM
                                                                                               oo    en
                                                                                                     io
                                                                                               o
                                                                                               i-H    CO
                                                                                               en
                                                                                               10
                                                                                               .-H    OO
                                                                                               O
                                                                                               o
                                                                                                     a
                                                                                                     oo
                                          504.1-18

-------
                    C7>
                    =1
 CO

 z
 o
 o
           5
           o
           CM
           
           O
CO
o
LU
o;
a.
o

                    CO
                    
                    >
                    o
                    u
                    O)
                            =«=

                             CO
                             O)
                            cc:
                                   to
CO

1—I

o
                        CO
                        CM
                              O
                              CO
                              CM
                                         oo
                               0
LO
o
CO
CM
o
CO
0
CO
CM
O
CO
CM
CM
CM
0
en
m
o
o
0
o
o
o
CM
*
0
                                          co
            01
            O
            CO
                  en
                  r^
                  en
                  oo
                  co
                                                co

                                                o
                                                            CD
                                                      CM
                                                        •

                                                      O
                              •<*•

                              CM

                              O
CM
CT>
CM
OO
IO
      
        o
        o
        CO
       o;
                                                                             n3
                                                                             CD
                                                                        UJ
                                                                        o
                                          O
                                          I—
                                          oo
                                                                                         O)
                                                                              Q.
                                                                              oo
                                                                                    a:
                                                                                    UJ    Q
                                                                                    >    oo
                                                                                    o    a:

CL
CC
Q
O_
CO
ex

CO
1 1


=**=
d)

CO

••-
Q
CO
Q£
en
00
en
o
0
00
I-H'
f~-
O
0
o
r-H
f
0





•^



LO
oo
CD
o
LO
en
o
o
>o
00
o
f






CM



O
O
i-H
cn
o
r— t
°
oo
IO
o
1 '
o





CO



t--. cn
en CM
0 0
«*• o
en CM
oo cn
0 0
0 0
LO •**•
LO CM
O r-H
i-H i-H
0 0





«3- LO



Cxi
IO
0
0
LO
CO
00
o
0
CM
OO
r-H
*""*
0





VO



r--.
r-H
°
0
10
o
•-H
0
1 —
CO
r-H
O





1 —



CO
o
o
en
en
o
o
oo
CM
i-H
r— t
O





§
CO
£


LU
o

Q
1—
/ r\
O
0
o
o
o
0
o
r-H
o
o
o
o
1 — 1
o
— 1
CD
=i

n
>
CO
— 1,

•*s
Q.
CO
0
I-H
en
F-H
en

oo
CM
I-H
i-H

rv
LU
5»
O
O
LU
O£

55
                                                                                         00
                                                                                         en
                                                                                                    00
                                                                                                    oo
                                                                                                   CO
                                                                                             •-H    O
                                                                                                   a
                                                                                                   oo
                                                                                                   a:
                                         504.1-19

-------
           TABLE 4.  INTERLABORATORY STUDY OF METHOD 504 REGRESSION
                     EQUATIONS FOR RECOVERY AND PRECISION*
                                                       1,2-Dibromo-
Vlater Type    	1.2-Dibromoethane	3-chloropropane	
Applicable Cone. Range     (0.05 - 6.68) M9/L          (0.05 - 6.40) /ig/L

Reagent Water
Single-Analyst Precision   SR = 0.041X -H 0.004         SR = 0.065X + 0.000
Overall Precision          S  = 0.075X + 0.008         S  = 0.143X - 0.000
Recovery                   X  = 1.072C - 0.006         X  = 0.987C - 0.000

Ground Water
Single-Analyst Precision   SR = 0.046X + 0.002         SR = 0.076X - 0.000
Overall Precision          S  = 0.102X + 0.006         S  = 0.160X + 0.006
Recovery                   X  = 1.077C - 0.001         X  = 0.972C + 0.007

X s Mean  recovery
C « True  value  for the  concentration
*   No  interlaboratory  method validation data is  available  for
    1,2,3-Trichloropropane using  Method  504,  Revision 3.0.
                                    504.1-20

-------
METHOD 505.  ANALYSIS OF ORGANOHALIDE PESTICIDES AND
 COMMERCIAL POLYCHLORINATED BIPHENYL (PCB) PRODUCTS
  IN WATER  BY MICROEXTRACTION AND GAS CHROMATOGRAPHY
                    Revision 2.1




             Edited  by J.W.  Munch  (1995)




  T.  W.  Winfield - Method 505, Revision 1.0 (1986)

  T.  W.  Winfield - Method 505, Revision 2.0 (1989)
       NATIONAL  EXPOSURE RESEARCH LABORATORY
        OFFICE OF RESEARCH AND DEVELOPMENT
       U.S. ENVIRONMENTAL PROTECTION AGENCY
              CINCINNATI,  OHIO  45268
                      505-1

-------
                                   METHOD 505
  ANALYSIS OF ORGANOHALIDE  PESTICIDES  AND  COMMERCIAL  POLYCHLORINATED  BIPHENYL
        (PCB) PRODUCTS IN WATER BY MICROEXTRACTION AND GAS CHROMATOGRAPHY

1.   SCOPE AND APPLICATION

     1.1  This method (1,2,3) is applicable to the determination of the
          following analytes in finished drinking water, drinking water during
          intermediate stages of treatment, and the raw source water:
               Analvte

               Alachlor
               Aldrin
               Atrazine
               Chlordane
               alpha-Chlorodane
               gamma-Chlorodane
               Dieldrin
               Endrin
               Heptachlor
               Heptachlor Epoxide
               Hexachlorobenzene
               Hexachlorocyclopentadi ene
               Lindane
               Methoxychlor
               cis-Nonachlor
               trans-Nonachlor
               Simazine
               Toxaphene
               Aroclor 1016
               Aroclor 1221
               Aroclor 1232
               Aroclor 1242
               Aroclor 1248
               Aroclor 1254
               Aroclor 1260
                               Chemical  Abstract Service
                                    Registry Number
                                      5972-
                                       309-
                                      1912-
                                        57-
                                      5103-
                                      5103-
                                        60-
                                        72-
                                        76-
                                      1024-
                                       118-
                                        77-
                                        58-
                                        72-

                                     39765-
                                       122-
                                      8001-
                                     12674-
                                     11104-
                                     11141-
                                     53469-
                                     12672-
                                     11097-
                                     11096-
60-8
00-2
24-9
74-9
71-9
74-2
57-1
20-8
44-8
57-3
74-1
74-4
89-9
•43-5

•80-5
•34-9
•35-2
•11-2
•28-2
-16-5
-21-9
-29-6
-69-1
-82-5
     1.2
     1.3
The analyst must demonstrate the applicability of the method by
collecting precision and accuracy data on fortified samples (i.e.,
groundwater, tap water) (4) and provide qualitative confirmation of
results by Gas Chromatography/Mass Spectrometry (GC/MS) (5), or by GC
analysis using dissimilar columns.

Method detection limits (MDL)  (6) for the above organohalides and
Aroclors have been experimentally determined  (Sect. 13.2).  Actual
detection limits are highly dependent upon the characteristics of the
gas chromatographic system used (e.g. column  type, age, and proper
conditioning; detector condition; and injector mode and condition).
                                      505-2

-------
      1.4   This  method  is  restricted  to  use  by  or  under  the  supervision of
           analysts  experienced  in  the use of GC and  in  the  interpretation of gas
           chromatograms.   Each  analyst  must demonstrate the  ability to generate
           acceptable results  with  this  method  using  the procedure described in
           Sect.  11.

2.    SUMMARY OF  METHOD

      2.1   Thirty-five mL  of sample are  extracted  with 2 ml of hexane.  One to
           two ill of the extract are  then injected into  a gas chromatograph
           equipped with a  linearized electron  capture detector for separation
           and analysis.  Analytes  are quantitated using procedural standard
           calibration (Sect 3.12).

      2.2   The extraction and  analysis time is  40  to  70 min per sample depending
           upon the analytes and the analytical  conditions chosen.  (See Sect
           6.9).

3.   DEFINITIONS

     3.1   LABORATORY DUPLICATES (LD1 and LD2)  - Two sample aliquots taken in
          the analytical laboratory and analyzed separately with identical
          procedures.   Analyses of LD1 and LD2  give a measure of the precision
          associated with laboratory procedures,  but not with sample collection
          preservation,  or storage procedures.                                  '

     3.2  FIELD DUPLICATES (FD1 and FD2) —  Two separate samples collected  at
          the same time and place under identical  circumstances  and  treated
          exactly the  same throughout field  and laboratory procedures.   Analyses
          of FD1 and FD2 give  a measure  of the  precision associated  with  sample
          collection,  preservation and storage, as well  as with  laboratory
          procedures.                                          ,,

     3.3  LABORATORY REAGENT BLANK (LRB) -  An  aliquot  of reagent water  that  is
          treated exactly  as a sample including exposure to  all  glassware,
          equipment, solvents, reagents, internal  standards,  and  surrogates that
          are  used with  other  samples.   The  LRB is used  to determine  if method
          analytes or other interferences  are present in the  laboratory
          environment,  the reagents,  or  the  apparatus.

     3.4  FIELD  REAGENT  BLANK  (FRB) - Reagent  water  placed  in a  sample
          container  in the  laboratory and treated  as  a sample in  all respects,
          including  exposure to  sampling site conditions,  storage, preservation
          and all analytical procedures.  The purpose of the  FRB  is to determine
          if method  analytes or  other interferences are  present  in the field
          environment.

     3.5   LABORATORY PERFORMANCE CHECK SOLUTION (LPC) — A solution of method
          analytes,  surrogate  compounds, and internal standards used to evaluate
          the performance  of the instrument system with  respect to a defined set
          of method criteria.
                                     505-3

-------
    3.6   LABORATORY  FORTIFIED  BLANK  (LFB)  — An  aliquot of reagent water to
          which  known quantities  of the method  analytes are added  in the
          laboratory.   The  LFB  is analyzed  exactly  like a  sample,  and  its
          purpose  is  to determine whether the methodology  is  in  control, and
          whether  the laboratory  is capable of  making  accurate and precise
          measurements at the required method detection limit.

    3.7   LABORATORY  FORTIFIED  SAMPLE MATRIX (LFM)  —  An aliquot of an environ-
          mental  sample to  which  known .quantities of the method  analytes are
          added  in the laboratory.  The  LFM is  analyzed exactly  like a sample,
          and its  purpose  is to determine whether the  sample  matrix contributes
          bias to  the analytical  results.   The  background  concentrations of the
          analytes in the  sample  matrix  must be determined in a  separate aliquot
          and the  measured  values in  the  LFM corrected for background  concentra-
          tions.

    3.8   STOCK  STANDARD SOLUTION —  A  concentrated solution  containing a  single
          certified standard that is  a  method  analyte, or  a concentrated
          solution of a single  analyte  prepared in  the laboratory  with an
          assayed  reference compound.  Stock standard  solutions  are used to
          prepare  primary  dilution standards.

    3.9   PRIMARY  DILUTION  STANDARD SOLUTION — A solution of several  analytes
          prepared in the  laboratory  from stock standard  solutions and diluted
          as needed to prepare  calibration  solutions  and  other needed  analyte :
          solutions.

    3.10 CALIBRATION STANDARD  (CAL)  — A solution  prepared from the  primary
          dilution standard solution  and stock standard  solutions  of  the
          internal standards and  surrogate  analytes.   The  CAL solutions are  used
          to calibrate the instrument response with respect to analyte
          concentration.

     3.11 QUALITY CONTROL SAMPLE  (QCS)  — A sample matrix containing  method
          analytes or a solution  of method  analytes in a water miscible solvent
          which  is used to fortify reagent  water or environmental  samples.   The
          QCS is obtained from a  source external to the  laboratory,  and is used
          to check laboratory performance with externally prepared test
          materials.

     3.12 PROCEDURAL STANDARD CALIBRATION —  A calibration method where  aqueous
          calibration standards  are prepared and processed (e.g. purged,
          extracted,  and/or derivatized) in exactly the  same manner as a sample.
          All steps in the process from addition of sampling preservatives
          through instrumental  analyses are included  in  the calibration.   Using
          procedural  standard calibration compensates for any inefficiencies in
          the processing procedure.

4.   INTERFERENCES

     4.1  Method  interferences may be caused by  contaminants in solvents,
          reagents, glassware and  other sample processing apparatus that lead to

                                      505-4

-------
 4.2
4.4
  discrete  artifacts  or elevated  baselines  in  gas  chromatograms    All
  reagents  and  apparatus must  be  routinely  demonstrated  to  be  free from
  interferences under the conditions  of  the analysis  by  running
  laboratory  reagent  blanks  as  described in Sect.  9.2.

  4.1.1  Glassware must be scrupulously  cleaned  (2).  Clean all glass-
        ware as  soon  as possible  after  use-by thoroughly rinsing with
        the  last solvent used  in  it.  Follow by washing with  hot water
        and detergent  and thorough rinsing with tap  and reagent water
        Drain  dry, and heat in an oven  or  muffle  furnace at 400°C for 1
        hr   Do not heat volumetric ware.   Thermally stable materials
        such as PCBs,  might not be eliminated by  this treatment
        Thorough rinsing with acetone may  be substituted for  the
        heating.  After  drying and cooling, seal  and store glassware in
        a clean environment to prevent  any  accumulation of dust or
        other  contaminants.   Store inverted or capped with aluminum
        T Q I I •

 4.1.2  The use of high purity reagents and solvents helps to minimize
        interference problems.  Purification of solvents by distilla-
        tion  in all-glass systems may be required.  WARNING:   When a
        solvent is purified, stabilizers put into the solvent by the
        manufacturer are removed  thus potentially making the  solvent
        hazardous.   Also, when a  solvent is purified, preservatives  put
        into  the solvent by  the manufacturer are removed thus
        potentially  reducing the  shelf-life.

 Interfering  contamination may occur  when a sample containing low
 concentrations of analytes  is analyzed  immediately following a  sample
 containing relatively high  concentrations  of  analytes.   Between-sample
 rinsing  of the sample syringe and associated  equipment  with  hexane  can
 minimize  sample  cross contamination.  After analysis of a  sample
 containing high  concentrations of analytes, one  or more  injections  of
 hexane should  be made to ensure  that  accurate values are obtained for
 the  next  sample.

 Matrix interferences  may be caused by contaminants that are coextract-
 ed from the  sample.   Also,  note that all the analytes listed  in the
 scope and  application  section are not resolved from  each other on any
 one  column,  i.e., one  analyte of  interest  may be  an  interferant for
 another analyte of interest.  The extent of matrix interferences will
 vary considerably from source to  source, depending upon the water
 sampled.  Cleanup of  sample extracts may be necessary   Analyte
 identifications should  be confirmed (Sect.  11.4).

 It is important that samples and  working standards be contained in the
same solvent   The solvent for working  standards must be the same as
the final  solvent used  in sample  preparation.   If this is not the
case, chromatographic comparability of standards to sample may be
affected.
                                505-5

-------
4.5  Caution must be taken in the determination of endrin since it has been
     reported that the splitless injector may cause endrin degradation (7).
     The analyst should be alerted to this possible interference resulting
     in an erratic response for endrin.

4.6  Variable amounts of pesticides and commercial PCB products from
     aqueous solutions adhere to glass surfaces.  It is recommended that
     sample transfers and glass surface contacts be minimized, and that
     adequate rinsing of glass surfaces be performed.

4.7  Aldrin, hexachlorocyclopentadiene and methoxychlor are rapidly
     oxidized by chlorine.  Dechlorination with sodium thiosulfate at time
     of collection will stop further oxidation of these compounds.

4.8  WARNING:  An interfering, erratic peak has been observed within the
     retention window of heptachlor during many analyses of reagent, tap,
     and groundwater.  It appears to be related to dibutyl phthalate;
     however, the specific source has not yet been definitively determined.
     The observed magnitude and character of this peak randomly varies in
     numerical value from successive injections made from the same vial.

SAFETY

5,1  The toxicity and carcinogenicity of chemicals used in this method have
     not been precisely defined; each chemical should be treated as a    -
     potential health hazard, and exposure to these chemicals should be
     minimized.  Each laboratory is responsible for maintaining awareness
     of OSHA regulations regarding safe handling of chemicals used in this
   -  method.  Additional references to laboratory safety are available
     (8-10) for the information of the analyst.

5.2  The following organohalides have been tentatively classified as known
     or suspected human or mammalian carcinogens:  aldrin, commercial PCB
     products, chlordane, dieldrin, heptachlor, hexachlorobenzene, and
     toxaphene.  Pure standard materials and stock standard solutions of
     these compounds should be handled in a hood or glovebox.

5.3  WARNING:  When a solvent is purified, stabilizers put into the solvent
     by the manufacturer are removed thus potentially making the solvent
     hazardous.
EQUIPMENT AND SUPPLIES  (All specifications are suggested.
are included for illustration only.)
Catalog numbers
6.1  SAMPLE CONTAINERS - 40-mL screw cap vials (Pierce #13075 or
     equivalent) each equipped with a size 24 cap with a flat, disc-like
     TFE facing backed with a polyethylene film/foam extrusion (Fisher
     #02-883-3F or equivalent).  Prior to use, wash vials and septa with
     detergent and rinse with tap and distilled water.  Allow the vials and
     septa to air dry at room temperature, place the vials in a 400°C oven
                                 505-6

-------
     for one hour, then remove and, allow to cool in an area known to be
     free of organics.

6.2  VIALS - auto sampler, screw cap with septa, 1.8 ml_, Varian
     #96-000099-00 or equivalent or any other autosampler vials not
     requiring more than 1.8.ml sample volumes.

6.3  AUTO SAMPLER - Hewlett-Packard 7671A, or equivalent.

6.4  MICRO SYRINGES -'lO and 100 /iL.

6.5  MICRO SYRINGE - 25 /iL with a 2-inch by 0.006-inch needle - Hamilton
     702N or equivalent.

6.6  PIPETTES - 2.0 and 5.0 ml transfer.

6.7  VOLUMETRIC FLASKS - 10 and 100 mL, glass stoppered.

6.8  STANDARD SOLUTION STORAGE CONTAINERS - 15-mL bottles with PTFE-lined
     screw caps.

6.9  GAS CHROMATOGRAPH — Analytical system complete^with temperature
     programmable GC and split/split!ess injector suitable for use with
     capillary columns and all required accessories including syringes,
     analytical columns, gases, a linearized electron capture detector and
     stripchart recorder.  A data system is recommended for measuring peak
     areas.  Table 1 lists retention times observed for method analytes
     using the columns and analytical conditions described below!

     6.9.1  Three gas chromatographic columns are recommended.  Column 1
            (Sect. 6.9.2) should be used as the primary analytical column
            unless routinely occurring analytes are not adequately
            resolved.  Validation data presented in this method were
            obtained using this column.  Columns 2 and 3 are recommended
            for use as confirmatory columns when GC/MS confirmation is not
            available.  Alternative columns may be used in accordance with
            the provisions described in Sect. 9.4.

     6.9.2  Column 1 (Primary Column) - 0.32 mm ID x 30 M long fused silica
            capillary with chemically bonded methyl polysiloxane phase
            (DB-1, 1.0 urn film, or equivalent).  Helium carrier gas flow is
            about 25 cm/sec linear velocity, measured at 180° with. 9 psi
            column head pressure.  The oven temperature is programmed from
            180°C to 260°C at 4°C/min and held at 260°C until all expected
            compounds have eluted.  Injector temperature:. 200°C.
            Splitless Mode:  0.5 min.  Detector temperature:  290°C.
            Sample chromatograms for selected pesticides are presented in
            Figures 1 and 2.  Chromatograms of the Aroclors, toxaphene, and
            technical chlordane are presented in Figures 3 through 11.

     6.9.3  Column 2 (alternative column 1) - 0.32mm ID x 30 M long fused
            silica capillary with a 1:1 mixed phase of dimethyl silicone

                                 505-7

-------
             and  polyethylene  glycol  (Durawax-DX3,  0.25/zm  film,  or
             equivalent).   Helium carrier  gas  flow  is  about  25 cm/sec  linear
             velocity  and  oven temperature is  programmed from 100°C to 210°C
             at 8°C/min, and held at  210°C until  all expected compounds have
             eluted.   Then the post temperature  is  programmed to 240°C at
             8°C/min for 5 min.

     6.9.4   Column 3  (alternative column  2) - 0.32mm  ID x 25 M  long fused
             silica capillary  with chemically  bonded 50:50 Methyl-Phenyl
             silicone  (OV-17,  1.5/zm film thickness, or equivalent).  Helium
             carrier gas flow  is  about  40  cm/sec  linear velocity and oven
             temperature is programmed  from 100°C to 260°C at 4°C/min  and
             held  at 260°C until  all  expected  compounds have eluted.

REAGENTS AND STANDARDS -  - WARNING:  When a solvent is purified,
stabilizers  put  into  the  solvent by  the manufacturer  are  removed thus
potentially  making the solvent hazardous.  Also, when a solvent is
purified, preservatives put into the solvent  by  the manufacturer are
removed thus potentially  making  the  shelf-life  short.

7.1  REAGENTS

     7.1.1   Hexane extraction solvent  - UV Grade,  Burdick and Jackson #216
             or equivalent.

     7.1.2   Methyl alcohol -  ACS Reagent  Grade,  demonstrated to be free of
             analytes.

     7.1.3   Sodium chloride,  NaCl -  ACS Reagent  Grade - For pretreatment
             before use, pulverize a  batch  of  NaCl  and place in  a muffle
             furnace at room temperature.   Increase the temperature to 400°C
             and hold  for  30 min.  Store in a  glass (not plastic) bottle to
             avoid phthalate contamination..

     7.1.4   Sodium thiosulfate,  Na?S203, ACS Reagent Grade—For  preparation
             of solution (0.04 g/mL), mix  1 g  of  Na2S203 with reagent water
             and bring to  25-mL volume  in  a volumetric flask.  Verify the
             stability of  this solution and replace as necessary.

7.2  REAGENT WATER -  Reagent  water is  defined as water free of  interference
     when employed in the  procedure  described herein.

     7.2.1   A Millipore Super-Q  Water  System  or  its equivalent  may be used
             to generate deionized reagent  water.

     7.2.2   Test reagent water each  day it is used by analyzing  it
             according to Sect. 9.2.

7.3  STOCK STANDARD SOLUTIONS -  These  solutions may be obtained  as
     certified solutions or prepared from  pure standard materials using the
     following procedures:


                                 505-8

-------
           7.3.1   Prepare  stock standard  solutions  (5000  /zg/mL)  by  accurately
                  weighing about 0.0500 g of pure material.   Dissolve  the
                  material  in  methanol and dilute to  volume  in  a 10-mL volumetric
                  flask.   Larger volumes  can be  used  at the  convenience of  the
                  analyst.   When compound purity is assayed  to  be 96%  or greater,
                  the weight can be  used  without correction  to  calculate the
                  concentration of the stock standard.  Commercially prepared
                  stock  standards can be  used  at any  concentration  if  they  are
                  certified  by the manufacturer  or by an  independent source.

           7.3.2   Transfer the stock standard  solutions into  Teflon-sealed
                  screw-cap  bottles.  Store  at 4°C and protect  from light.  Stock
                  standard solutions should  be checked frequently for  signs of
                  degradation  or evaporation,  especially  just prior to preparing
                  calibration  standards from them.

           7.3.3   Stock  standard solutions must  be replaced  after six  months, or
                  sooner if  comparison with  check standards  indicates  a  problem.

     7.4   PRIMARY DILUTION  STANDARD SOLUTIONS — Use stock  standard solutions to
           prepare primary dilution  standard solutions that contain the  analytes
           in methanol.  The primary dilution  standards should be prepared  at
           concentrations  that  can be easily diluted  to prepare  aqueous  calibra-
           tion standards  (Sect. 10.2.1)   that will bracket the working concentra-
           tion range.  Store  the primary dilution standard solutions  with
           minimal headspace and check frequently for signs of deterioration or
           evaporation, especially just before preparing  calibration standards.
           The storage time described for stock  standard  solutions in  Sect. 7.3.3
           also applies to primary dilution  standard  solutions.    Note:  Primary
           dilution standards  for toxaphene, chlordane and each  of the Aroclors
           must be prepared  individually.

8.   SAMPLE COLLECTION. PRESERVATION.  AND STORAGE

     8.1   SAMPLE COLLECTION

          8.1.1  Collect all samples in  40-mL bottles into which 3 mg  of sodium
                 thiosulfate crystals  have been added to the empty bottles just
                 prior to shipping to  the sampling  site.  Alternately,  75 #L Of
                 freshly prepared sodium thiosulfate  solution (0.04 g/mL)  may be
                 added to empty 40-mL  bottles just  prior to sample collection.

          8.1.2  When sampling from a  water tap, open the tap and allow the
                 system to flush until  the water temperature has stabilized
                 (usually about 10 min).   Adjust the  flow to about 500 mL/min
                 and collect samples from the flowing stream.

          8.1.3  When sampling from a  well,  fill a  wide-mouth bottle or beaker
                 with sample,  and carefully fill 40-mL sample bottles.

     8.2  SAMPLE PRESERVATION


                                     505-9

-------
          8.2.1  The  samples must  be chilled to 4°C at the time of collection
                 and  maintained  at that temperature until the analyst is
                 prepared  for the  extraction process.  Field samples that will
                 not  be received at the laboratory on the day of collection must
                 be packaged for shipment with sufficient ice to insure that
                 they will be maintained at 4°C until arrival at the laboratory.

     8.3  SAMPLE STORAGE

          8.3.1  Store samples and extracts at 4°C until extraction and
                 analysis.

          8.3.2  Extract all samples as soon as possible after collection.
                 Results of holding time studies suggest that all analytes with
                 the  possible exception of heptachlor were adequately stable for
                 14 days when stored under these conditions.  In general,
                 heptachlor showed inconsistent results.  If heptachlor is to be
                 determined, samples should be extracted within 7 days of
                 collection.  The maximum holding time for all other analytes is
                 14 days.  Analyte stability may be affected by the matrix;
                 therefore, the analyst should verify that the preservation
                 technique is applicable to the samples under study.

          8.3.3  Because of the potential for extract volume loss due to
                 evaporation, it is recommended that extracts be analyzed
                 immediately after preparation.  If this is not possible,
                 extracts may be stored at 4°C or less in tightly capped vials
                 (Sect. 6.2) for up to 24 hours.

9.   QUALITY CONTROL

     9.1  Minimum quality control (QC) requirements are initial  demonstration of
          laboratory  capability,  analysis of laboratory reagent blanks (LRB),
          laboratory  fortified blanks (LFB),  laboratory fortified sample matrix
          (LFM), and quality control  samples  (QCS).  A MDL for each analyte must
          also be determined.

     9.2  Laboratory Reagent Blanks.   Before  processing any samples,  the analyst
          must demonstrate that all glassware and reagent interferences are
          under control.   Each time a set of  samples is extracted or reagents
          are changed, an LRB must be analyzed.  If within the retention time
          window of any analyte the LRB produces  a  peak that would prevent the
          determination of that analyte,  determine  the source of contamination
          and eliminate the interference before processing samples.

     9.3  Initial  Demonstration of Capability

          9.3.1  Select a representative fortified  concentration (about 10 times
                 MDL or at a concentration in the middle of the  calibration
                 range established in Section 10)  for each analyte.   Prepare a
                 standard concentrate containing  each analyte at 1000 times the
                 selected concentration.   With a  syringe,  add 35 fjl of the

                                     505-10

-------
             concentrate to each of four to seven 35 ml aliquots of reagent
             water,  and analyze each aliquot according to procedures
             beginning in Sect. 11.

      9.3.2   For each analyte the mean recovery value for these samples  must
             fall  in the range of ± 30%  of the fortified amount.   The RSD
             for these measurements  must be 20% or less.   For those
             compounds that meet the acceptance criteria,  performance  is
             considered acceptable.   For those  compounds  that fail  these
             criteria,  this procedure must  be repeated using  fresh  replicate
             samples until  satisfactory performance has been  demonstrated.

      9.3.3   For each analyte,  determine the MDL.   Prepare a  minimum of  7
             LFBs  at a low  concentration.   Use  calibration data obtained in
             Section 10 to  estimate  a concentration for each  analyte that
             will  produce a peak with a 3-5 times  signal  to noise response.
             Extract and analyze each replicate according  to  Sections  11 and
             12.   It is recommended  that these  LFBs be prepared and analyzed
             over  a  period  of several  days,  so  that day to day  precision is
             reflected  in the precision measurements.   Calculate mean
             recovery and standard deviation  for  each  analyte.  Use  the
             equation given in  Sect.13  to calculate the MDL.

      9.3.4   The  initial  demonstration  of capability  is used  primarily to
             preclude a laboratory from analyzing  unknown  samples via a  new,
             unfamiliar method  prior  to  obtaining  some  experience with it.
             It  is expected that  as  laboratory  personnel gain experience
             with  this  method the quality of data  will  improve  beyond those
             required  here.

9.4  The analyst  is  permitted  to modify GC columns, GC conditions,
     concentration  techniques  (i.e.  evaporation techniques),  internal
     standards or surrogate  compounds.  Each time  such method modifications
     are made, the  analyst must repeat  the procedures  in  Sect. 9.3.

9.5  Assessing Laboratory  Performance - Laboratory Fortified Blank  (LFB)

     9.5.1  The laboratory must analyze at least one laboratory fortified
            blank (LFB) per sample set  (all samples extracted within a 24-h
            period).   If the sample  set contains more than 20 samples,
            analyze one LFB for every 20 samples.  Ideally the fortifying
            concentration of each analyte in the LFB sample should  be  the
            same as that selected in Sect.  9.3.1.  Calculate  accuracy  as
            percent recovery (X,-).   If the  recovery of any analyte  falls
            outside the control limits  (see Sect. 9.5.2), that analyte is
            judged out of control,  and the  source of the  problem should be
            identified and  resolved before  continuing analyses.  Because
            this LFB is prepared and analyzed in the same way as a
            calibration verification standard,  it can be  used to satisfy
            the calibration requirement in  Sect.10.2.3.
                                505-11

-------
            Note:  It is suggested that one multi-component analyte
            (toxaphene, chlordane or an Aroclor)  LFB also be analyzed with
            each sample set.   By selecting a different multi-component
            analyte for this  LFB each work shift,   LFB data can be obtained
            for all of these  analytes over the course of several  days

     9.5.2  Until  sufficient  data become available from within their own
            laboratory, usually a minimum of results from 20 to 30
            analyses,  the laboratory may assess laboratory performance
            against the control limits in Sect. 9.3.2 that are derived from
            the data in Table 2.  When sufficient  internal performance data
            becomes available, develop control limits from the mean percent
            recovery (X) and  standard deviation (S) of the percent
            recovery.   These  data are used to establish upper and lower
            control limits as follows:                               ;

                    UPPER CONTROL LIMIT  =  X + 3S
                    LOWER CONTROL LIMIT  =  X - 3S

            After each five to ten new recovery measurements, new control.
            limits should be  calculated using only the most recent 20-30
            data points.  These calculated control limits should not exceed
            the fixed limits  in Sect. 9.3.2.

     9.5.3  It is recommended that the laboratory periodically determine
            and document its  detection limit capabilities for analytes of
            interest.   CAUTION:  No attempts to establish low detection
            limits should be  made before instrument optimization and
            adequate conditioning of both the column and the GC system.
            Conditioning includes the processing of LFB and LFM samples
            containing moderate concentration levels of these analytes.

     9.5.4  At least quarterly the laboratory should analyze quality
            control samples (QCS).  If acceptance criteria are not met,
            corrective action should be taken and documented.

9.6  Assessing Analyte Recovery - Laboratory Fortified Sample Matrix (LFM)

     9.6.1  The laboratory must add a known concentration to a minimum of
            10% of the routine samples or one LFM per set, whichever is
            greater.  The fortified concentration should not be less than
            the background concentration of the sample selected for
            fortification.  Ideally the LFM concentration should be the
            same as that used for the LFB (Sect. 9.3.1).  Periodically,
            samples from all  routine sample sources should be fortified.

     9.6.2  Calculate the percent recovery  (R,-) for each analyte,  corrected
            for background concentrations .measured  in the unfortified
            sample.

     9.6.3  If the recovery of any such analyte falls outside the range of
            ± 35% of the fortified amount,  and the  laboratory performance

                                505-12

-------
9.7
                 for that analyte is shown to be in control (Sect. 9.5), the
                 recovery problem encountered with the dosed sample is judged to
                 be matrix related, not system related.  The result for that
                 analyte in the unfortified sample is labeled suspect/matrix to
                 inform the data user that the results are suspect due to matrix
                 effects.

          The laboratory may adopt additional quality control  practices for use
          with this method.  The specific practices that are most productive
          depend upon the needs of the laboratory and the nature of the samples.
          For example, field or laboratory duplicates may be analyzed to assess
          the precision of the environmental  measurements or field reagent
          blanks may be used to assess contamination of samples under site
          conditions, transportation and storage.

10.   CALIBRATION AND STANDARDIZATION

     10.1 Establish GC operating parameters equivalent to those indicated in
          Sect.  6.9.  WARNING:   Endrin is easily degraded in the injection port
          if the injection port or front of the column is dirty.  This is the
          result of buildup of high boiling residue from sample injection.
          Check for degradation problems daily by injecting a  mid-level  standard
          containing only endrin.   Look for the degradation products of endrin
          (endrin ketone and endrin aldehyde).  If degradation of endrin exceeds
          20%,  take corrective action before  proceeding with calibration.      :
          Calculate percent breakdown as follows:

     Total  endrin degradation peak area fendrin aldehyde + endrin ketone)    xlOO%
     Total  endrin peak area (endrin + endrin  aldehyde + endrin ketone)

     10.2 At least three calibration standards are needed;  five are
          recommended.   Guidance on the number of standards is as follows:  A
          minimum of three calibration standards  are  required  to calibrate a
          range  of a factor of 20  in concentration.   For a factor of 50  use at
          least  four standards,  and for a factor  of 100 at least five
          standards.  The lowest standard should  represent analyte
          concentrations near,  but above,  their respective MDLs.   The
          remaining standards should bracket  the  analyte concentrations
          expected in the sample extracts,  or should  define the working  range
          of the detector.

          10.2.1  To prepare  a calibration  standard  (CAL),  add  an  appropriate
                 volume  of a  secondary dilution  standard to a  35  ml aliquot of
                 reagent water  in  a  40-mL  bottle.   Do  not  add  less than  20 /xl_
                 of an  alcoholic standard  to  the  reagent water.   Use a 25-/iL
                 micro  syringe  and  rapidly  inject the  methanolic  standard  into
                 the middle  point  of the  water volume.   Remove  the needle  as
                 quickly as  possible  after  injection.   Mix  by  inverting  and
                 shaking the  capped  bottle  several  times.   Aqueous standards
                 must be prepared  fresh  daily.
                              505-13

-------
     10.2.2 Starting with the standard of lowest concentration, prepare,
            extract, and analyze each calibration standard beginning with
            Sect. 11.2 and tabulate peak height or area response versus
            the concentration in the standard.  The results are to be
            used to prepare a calibration curve for each compound by
            plotting the peak height or area response versus the concen-
            tration.  Alternatively, if the ratio of concentration to
            response (calibration factor) is a constant over the working
            range (20% RSD or less), linearity to the origin can be
            assumed and the average ratio or calibration factor can be
            used in place of a calibration curve.  NOTE: Toxaphene,
            chlordane, and Aroclor standards must be injected separately.
            See Section 12.2 for information on quantitation of multi-
            component analytes.

     10.2.3 The working calibration curve or calibration factor must be
            verified on each working day by the measurement of a minimum
            of two calibration check standards, one at the beginning and
            one at the end of the analysis day.  These check standards
            should be at two different concentration levels to verify the
            calibration curve.  For extended periods of analysis (greater
            than 8 hrs.), it is strongly recommended that check standards
            be interspersed with samples at regular intervals during the
            course of the analyses.  If the response for any analyte
            varies from the predicted response by more than the criteria :
            in Sect. 9.3.2, the test must be repeated using a fresh
            calibration standard.  If the results still do not agree,
            generate a new calibration curve.  For those analytes that
            failed the calibration verification, results from field
            samples analyzed since the last passing calibration should be
            considered suspect.  Reanalyze sample extracts for these
            analytes after acceptable calibration is restored.

            Note: It is suggested that a calibration verification
            standard of one multi-component analyte, either chlordane,
            toxaphene or an Aroclor also be analyzed each work shift.  By
            selecting a different multi-component analyte for this
            calibration verification each work shift, continuing
            calibration data can be obtained for all of these analytes
            over the course of several  days.

10.3 INSTRUMENT PERFORMANCE - Check the performance of the entire
     analytical system daily using data gathered from analyses of
     laboratory reagent blanks (LRB),  (CAL), and laboratory duplicate
     samples (LD1 and LD2).

     10.3.1 Significant peak tailing in excess of that shown for the
            target compounds in the method chromatograms (Figures 1-11)
            must be corrected.  Tailing problems are generally traceable
            to active sites on the GC column, improper column installa-
            tion, or operation of the detector.f


                               505-14

-------
          10.3.2 Check the precision between replicate analyses.  Poor
                 precision is generally traceable to pneumatic leaks,
                 especially at the injection port.  If the GC system is
                 apparently performing acceptably but with decreased
                 sensitivity, it may be necessary to generate a new curve or
                 set of calibration factors to verify the decreased responses
                 before searching for the source of the problem.

          10.3.3 Observed relative area responses of endrin (Sect. 10.1) must
                 meet the following general criteria:

                 10.3.3.1 The breakdown of endrin into its aldo and keto forms
                          must be adequately consistent during a period in
                          which a series of analyses is made.   Equivalent
                          relative amounts of breakdown should be demonstrated
                          in the LRB, LFB, CAL and QCS.  Consistent break-
                          down resulting in these analyses would suggest that
                          the breakdown occurred in the instrument system and
                          that the methodology is in control.

                 10.3.3.2 Analyses of laboratory fortified matrix (LFM)
                          samples must also be adequately consistent after
                          corrections for potential  background concentrations
                          are made.

11.   PROCEDURE

     11.1 SAMPLE PREPARATION

          11.1.1 Remove samples from storage and allow them to equilibrate to
                 room temperature. .

          11.1.2 Remove the container caps.   Withdraw and discard a 5 mL
                 volume using a 10-mL graduated cylinder.   Replace the
                 container caps and  weigh the containers  with  contents to the
                 nearest 0.1  g and  record these weights for subsequent sample
                 volume determinations  (Sect.  11.3).

     11.2 EXTRACTION AND ANALYSIS

          11.2.1 Remove the container cap of each sample,  and  add 6 g NaCl
                 (Sect.  7.1.3)  to the sample bottle.   Using a  class A,  TD
                 transfer or  automatic  dispensing pipet,  add 2.0  mL of hexane.
                 Recap  and shake  vigorously  by  hand  for 1  min.   Invert the
                 bottle and allow the water  and hexane  phases  to  separate.

          11.2.2 Remove the cap and  carefully- transfer  approximately 0.5 mL  of
                 hexane layer into an autosampler vial  using a disposable
                 glass  pipet.

          11..2.3 Transfer the remaining  hexane  phase, being careful  not to
                 include  any  of the  water phase,  into a second autosampler

                                    505-15

-------
            vial.  Reserve this second vial  at 4°C for reanalysis if
            necessary.

     11.2.4 Transfer the first sample vial  to an autosampler set up to
            inject 1-2 fil portions into the gas chromatograph for
            analysis (See Sect. 6.9 for GC conditions).  Alternately,  1-2
            fjl portions of samples, blanks,  and standards may be manually
            injected, although an autosampler is strongly recommended.

11.3 DETERMINATION OF SAMPLE VOLUME IN, BOTTLES NOT CALIBRATED

     11.3.1 Discard the remaining sample/hexane mixture from the sample
            bottle.  Shake off the remaining few drops using short, brisk
            wrist movements.

     11.3.2 Reweigh the empty container with original cap and calculate
          • the net weight of sample by difference to the nearest 0.1  g
            (Sect. 11.1.2 minus Sect. 11.3.2).  This net weight (in
            grams) is equivalent to the volume (in mL) of water extracted
            (Sect. 12.4).  Alternatively, by using 40-mL bottles precali-
            brated at 35-mL levels, the gravimetric steps can be omitted,
            thus increasing the speed and ease of this extraction
            process.

11.4 IDENTIFICATION OF ANALYTES

     11.4.1 Identify a sample component by comparison of its retention
            time to the retention time of a reference chromatogram.  If
            the retention time of an unknown compound corresponds, within
            limits, to the retention time of a standard compound, then
            identification is considered positive.

     11.4.2 The width of the retention time window used to make identifi-
            cations should be based upon measurements of actual retention
            time variations of standards over the course of a day.  Three
            times the standard deviation of a retention time can be used
            to calculate a suggested window size for a compound.
            However, the experience of the analyst should weigh heavily
            in the interpretation of chromatograms.

     11.4.3 Identification requires expert judgement when sample compo-
            nents are not resolved chromatographically.  When peaks
            obviously represent more than one sample component  (i.e.,
            broadened peak with shoulder(s) or valley between two or more
            maxima), or any time doubt exists over the identification of
            a peak on a chromatogram, appropriate alternative techniques
            to help confirm peak-identification need be employed.  For
            example, more positive identification may be made by the use
            of an alternative detector which operates on a
            chemical/physical principle different from that originally
            used, e.g.-, mass spectrometry, or the use of a second
                               505-16

-------
                 chromatography column.  Suggested alternative columns are
                 described in Sect. 6.9.

                 Note: Identify multi-component analytes by comparison of the
                 sample chromatogram to the corresponding calibration standard
     •   .         chromatograms of chlordane, toxaphene and the Aroclors.
                 Identification of multi-component analytes is made by pattern
                 recognition, in which the experience of the analyst is an
                 important factor.

12.   DATA ANALYSIS AND CALCULATIONS

     12.1 Identify the organohalides in the sample chromatogram by comparing
          the retention time of the suspect peak to retention times generated
          by the calibration standards and the laboratory fortified blanks.
         . Identify the multicomponent compounds using all peaks that are
          characteristic of the specific compound from chromatograms generated
          with individual  standards.

     12.2 To quantitate multi-component analytes,  one, of the following methods
          should be used.
          Option 1- Calculate an average response factor or linear regression
      ;    equation for each multi-component analyte using the combined area  of
          all  the component peaks in each of the calibration standard
          chromatograms.                                                       :
          Option 2- Calculate an average response factor or linear regression
          equation for each multi-component analyte using the combined areas
          of 3-6 of the most  intense and reproducible peaks in each of the
          calibration  standard chromatograms.

          When quantifying multi-component  analytes in samples,  the analyst
          should use caution  to include only those peaks from the  sample  that
          are  attributable to the multi-component  analyte.   Option 1  should
          not  be used  if there are significant interference peaks  within  the
          chlordane, Aroclor  or toxaphene pattern.

     12.3  Use  the multi-point calibration curve or calibration factor (Sect.
          10.2.3)  to directly calculate the uncorrected  concentration (Ci) of
          each analyte in  the sample (e.g.,  calibration  factor x response).
          Do not use the daily calibration  standard to quantitate  method
          analytes  in  samples.   If any  analyte response  exceeds the
          calibration  range,  dilute  the extract and reanalyze.

     12.4  Calculate  the  sample volume  (Vs)  as  equal  to the  net sample weight:

          Vs = gross weight  (Sect.  11.1,2)  -  bottle tare  (Sect. 11.3.2).

     12.5  Calculate  the corrected  sample  concentration  as:

          Concentration,   /ig/L  =  35(C-)
                                  (V.)


                                    505-17

-------
     12.6 Results should be reported with an appropriate number of significant
          figures.  Experience indicates that three significant figures may be
          used for concentrations above 99 /*g/L, two significant figures for
          concentrations between 1-99 M9/U and 1 significant figure for lower
          concentrations.

13.   METHOD PERFORMANCE

     13.1 Single laboratory (NERL-Cincinnati) accuracy and precision at
          several concentrations in reagent, ground, and tap water matrices
          are presented in Table 2.  These results were obtained from data
          generated with a DB-1 column, and with quantitation Option 2 as
          described in Section 12.2.


     13.2 With these data, the method detection limits (MDL) in Table 2 were
          calculated using the formula:

                          MDL = S t(n.1(1.alpha _ 0_99)

                 where:

                          tcn-i 1-aipha = o 99x = Student's  t  value  for  the  99%
                          confidence"level with n-1 degrees of freedom
                          n = number of replicates                            :
                          S = standard deviation of replicate analyses.

     13.3 This method has been tested by 10 laboratories using reagent water
          and groundwater fortified at three concentration levels.  Single.
          operator precision, overall precision, and method accuracy were
          found to be directly related to the concentration of the analyte and
          virtually independent of the  sample matrix.  Linear equations to
          describe the relationships are presented  in Table 3.

14.   POLLUTION PREVENTION

     14.1 This method utilizes a microextraction procedure that requires the
          use of very small volumes of  hexane,  thus making this method safe
          for use by the laboratory analyst and harmless to the environment.
          For information concerning pollution  prevention that may be
          applicable to laboratory operations,  consult "Less is Better:
          Laboratory Chemical Management for Waste  Reduction" available from
          the American Chemical Society's Department of  Government Relations,
          and Science Policy, 1155 16th Street  N.W., Washington, D.C. 20036.

15.   WASTE MANAGEMENT

     15.1 It is the laboratory's responsibility to  comply with all federal,
          state, and local regulations  governing the waste management,
          particularly the hazardous waste  identification rules and land
          disposal restrictions, and to protect the air, water, and land by
          minimizing and controlling all releases from fume hoods  and bench

                                    505-18

-------
          operations.  Also, compliance is required with any sewage discharge
          permits and regulations.  For further information on waste
          management, consult "The Waste Management Manual for Laboratory
          Personnel," also available from the American Chemical Society at the
          address in Sect. 14.1.

16.   REFERENCES

     1.   Glaze, W.W., Lin, C.C., Optimization of Liquid-Liquid Extraction
         Methods for Analysis of Organics in Water, EPA-600/S4-83-052, January
         1984.

     2.   Henderson,  J.E.,  Peyton,  G.R.  and Glaze,  W.H. (1976).   In
         "Identification and Analysis  of Organic Pollutants in Water"  (L H
         Keith  ed.),  pp.  105-111.   Ann  Arbor Sci.  Publ.,  Ann  Arbor,  Michigan.

     3.   Richard,  J.J.,  Junk,  G.A.,  "Liquid Extraction for Rapid  Determination
         of  Halomethanes in Water,"  Journal  AWWA,  69,  62,  January 1977.

     4.   "Handbook for Analytical  Quality Control  in  Water and  Wastewater
         Laboratories,"  EPA-600/4-79-019,  U.  S.  Environmental  Protection
         Agency,  Environmental Monitoring  and Support  Laboratory   Cincinnati
         Ohio,  45268, March 1979.                                     '       '

     5.   Munch,  J. W., "Method 525.2-Determination  of  Organic  Compounds  in
         Drinking  Water  by Liquid-Solid  Extraction  and Capillary  Column
         Chromatography/ Mass Spectrometry"  in Methods for  the  Determination
         of  Organic Compounds in Drinking  Water? Supplement 3   (iqgs)~   [JSEPA
         National  Exposure Research  Laboratory,  Cincinnati, Ohio  45268.

     6.   Glaser, J.A. et al., "Trace Analyses  for Wastewaters," Environmental
         Science and  Technology, 15, 1426  (1981).

     7.   Bellar, T.A., Stemmer,  P.,  Lichtenberg, J.J., "Evaluation of
         Capillary Systems  for the Analysis of Environmental Extracts  "
         EPA-600/S4-84-004, March 1984.

    8.   "Carcinogens-Working with Carcinogens," Department of Health,
         Education, and Welfare,  Public Health Service, Center for Disease
         Control, National  Institute of Occupational Safety and Health
         Publication No.  77-206,  August, 1977.

    9.   "OSHA Safety and Health Standards, General Industry," (29CFR1910)
        Occupational  Safety and Health Administration, OSHA 2206, (Revised
        January 1976).                                         .            '

   10.   "Safety in Academic Chemistry Laboratories,"  American Chemical
        Society Publication, Committee on Chemical Safety, 3rd Edition,  1979.
                                   505-19

-------
17.  TABLES. DIAGRAMS. FLOWCHARTS. AND VALIDATION DATA

                 TABLE 1. RETENTION TIMES FOR METHOD ANALYTES
                                               Retention Titne(a), Min
                                 Primary     Confirm. 1Confirm. 2
Hexachlorocyclopentadiene 5
Simazine 10
Atrazine 11
Hexachlorobenzene 11
Lindane 12
Alachlor 15
Heptachlor 15
Aldrin , 17
Heptachlor Epoxide 19
gamma-Chlordane 19
alpha-Chlordane 20
trans-Nonachlor 21
Dieldrin 22
Endrin 23
cis-Nonachlor 24
Methoxychlor 30
.5
.9
.2
.9
.3
.1
.9
.6
.0
.9
.9
.3
.1
.2
.3
.0


6
25
22
13
18
19
17
18
24
25
26
24
45
33
39
58
.8
.7
.6
.4
.4
.7
.5
.4
.6
.9
.6
.8
:i
.3
.0
.5
5.2
19.9
19.6
15.6
18.7
21.1
20.0
21.4
24.6
26.0
26.6
26.3
27.8
29.2
30.4
36.4
Primary(b)
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
Chlordane
Toxaphene
13
7.
11
11
14
19
23
15
21
.6
7,
.2
.2
.8
.1
.4
.1
.7
, 14
9.0
, 14
, 13
, 16
, 21
, 24
, 15
, 22
.8
3
.7
.6
.2
.9
.9
.9
.5
, 15.
15.9,
, 13.
, 14-
, 17-
, 23.
, 26.
, 20.
, 26.
2

6
7
1
4
7
1
7
, 16.
19.1,
, 15.
, 15.
, 17.
, 24.
, 28.
, 20.
, 27.
2,
24
2,
2,
7,
9,
2,
9,
2
17
.7
17
17
19
26
29
21

.7

.7
.7,
.8,
.7
.9,
.3




19.8
22.0

32.6


     Columns and analytical conditions are described in Sect. 6.9.2, 6.9.3,
     and 6.9.4.

     Column and conditions described  in Sect. 6.9.2.  More than one peak
     listed does not indicate the total number of peaks characteristic of the
     multi-component analyte.  Listed peaks  indicate only the ones chosen for
     summation in the quantification.
                                    505-20

-------
TABLE 2.  SINGLE LABORATORY ACCURACY, PRECISION AND METHOD DETECTION LIMITS
             (MDLS)  FOR ANALYTES  FROM REAGENT WATER,  GROUNDWATER,  AND TAP  WATER3

                                          Accuracy  and  Standard  Deviation  Data
Analvte
Alachlor
Aldrin
Atrazine

alpha-Chlordane

gamma-Chlordane

Chlordane

Dieldrin

Endrin

Heptachlor

Heptachlor Epoxide

Hexachl orobenzene

Hexachl orocycl operitadiene

Lindane

Methoxychlor

cis-Nonachlor

trans-Nonachlor

Simazine

6.2
Toxaphene

Aroclor 1016
MDL
ttd/L
0.225
0.007
2.4

0.006

0.012

0.14

0.012

0.063

0.003

0.004

0.002

0.13

0.003

0.96

0.027

0.011

6.8


1.0

0.08
Aroclor 1221 15.0
Aroclor 1232
Aroclor 1242
Aroclor 1248

Aroclor 1254

Aroclor 1260

0.48
0.31
0.102

0.102

0.189

Concen-
tration*
mil
0.50
0.05
5.0
20.0
0.06
0.35
0.06
0.35
0.17
3.4
0.10
3.6
0.10
3.6
0.032
1.2
0.04
1.4
0.003
0.09
0.15
0.35
0.03
1.2
2.10
7.03
0.06
0.45
0.06
0.35
25
60

10
80
1.0
180
3.9
4.7
3.6
3.4
1.8
1.7
2.0
1.8
Reagent Water
R* S.d
102
106
85
95
95
86
95
86
-
-
87
114
119
99
77
80
100
115
104
103
73
73
91
111
100
98
110
82
95
86
99
65

-
--
_
_
_
_
_
_
_
-
_
-
13.4
20.0
16.2
5.2
3.5
17.0
0.4
18.5
8.0
3.6
17.1
9.1
29.8
6.5
10.2
7.4
15.6
6.6
13.5
6.6
5.1
11.7
6.5
5.0
21.0
10.9
15.2
21.3
9.6
21.8
8.3
3.6

_
_
_
_
_
_
_
_
_
_•
_
_
Groundwater
R S.

86
95
86
83
94
86
95

_
67
94
94
100
37
71
90
103
91
101
87
69
88
109

—
101
93
83
94
97
59

_
_
_
_
_
_
_
_
_
..
_
_
	 -K 	
16.3
7.3
9.1
4.4
10.2
5.3
14.5

-.
10.1 .
8.6
20.2
11.3
6.8
9.8
14.2
6.9
10.9
4.4
5.1
4.8
7.7
3.4

	
7.2
18.3
7.1
17.2
9.2
18.0

_
_
	
_
„
_
--p
—
_
_
_
—
Tap
R


108
91
•J J,
85
91
83
91
105
95
92
81
106
85
200
106
112
81
100
88
191
109
103
93

_
93
87
73
86
102
67

110
114
97
92
86
96

84

85

88
Water
Sr
	 -R—

10.9
3 1
w • X
7.1
2.4
14.7
6 0
V * V
12.4
96
••/ • w
15.7
14.0
14.0
12.4
22.6
16.8
7.5
5.9
15.6
13.4
18.5
14.3
8.1
18.4

_
14.3
5.4
4.1
5.1
13.4


9.5
13.5
7.5
9.6
7.3
7.4

9.9

11.8

19.8
                                  505-21

-------
                             Table 2  (Continued)


aData corrected for amount detected in blank and represents the mean  of 5-8
samples.
       method detection limit in sample in M9/L;  calculated by multiplying
standard deviation (S) times the students' t value appropriate for a 99%
confidence level and a standard deviation estimate with n-1 degrees of
freedom.

CR = average percent recovery.

dSD - Standard deviation  about  percent recovery.
  K

*  Refers to  concentration levels used to generate R and SR data for the three
types of water Matrices, not for MDL  determinations.

-  No  analyses conducted.
                                     505-22

-------
in
o
in
 I
o
t—t

i
o
o
o

to

o
I-H
t—
co
o
I—I
CO
CO
CD
0
CO
•i—
(J <•— »
0) —1
Q. oo cn
IH "^
(U
o




+J
CO
>> u
r— CO
ro
C C
CO CO --.
i 	 r- U>
cn u 23.
C CD **->
•t— S-
CO Q-



co
« X
u s- cr
ro eo --.
s- > cn
3 0 =1
0 O 	
o cw
< o£



(U O>
-Q CTJ *•*»»
ro Qi —I
0 -»
•r- . 0>
•^ y. =*•
^3u £« ^^*
a. o














s- •
0)
•M
CD
E
ro
s-
ro
a.

CO
CM
CM
'|S
O
O






1 — 1
CM
i-H
IX
O
O
0
0




1**^
en
o
CM
CM
I-H



O
en
in
*^
i
10
o
ro















CU
c
•r—
tsl
re
^
+J

co en
CM O
0 CD




i*** en
ro ro

o --H
i i
•— i -a-
o o
O 0






cu
£=
cu
N
c
ai
J3
0

O
r— O)
-C C
O re
re -o
x c:
CU •!-
in — i

VO
i— t
0
IX
in
o
o






o
I-H
0
1 X
1-^
r--
o
o




00
o
o
1
o




o
in
f^.
ro
i
o
in
o















^
0
r—
Jd
O
ro
f—
«=C

CM O
^3 C3
0 O
IJX I X
f-H VO
CM CM
0 O






•— I CM
0 0
o o
IX IX
|-~ i-H
O ro
•-1 0
o o




CM O
0 O
0 0
o o
CM IO
CZ) VO




•-H CM
• •
1 — 1 1 — 1
1 1
czi o
0 0













s_
o
r—
f-
o c
ro -i-
•4-> S-
Q. "O
CU i —
in  o
C— ) <-J
CM r«-
in CM
cn o
O fH



CM CO
• •
f—i r~- .
i i
»3- O
O I-H
0 O





0)

• f*—
X
o
a.
cu

c_
O
i — C
-C -f-
u s-
ro T3
•*-> r—
a. a>
cy -•-
rc a

CM
o
o
IX
IO
ro
o






i-H
ez>
o
'1
f—4
I-H
O




• i-H
O
.958C+0.
0



0
in

i^
i
o
i-H
O

















c
•f—
s-

c:
LU

o cn
CM i— t
0 O
ix ,x
m in
CM CM
0 O






CM IO
f-H O
O 0
'5 '3
^ oo
i-H O
o 

. in ro cn o O f-H 0 O o en in o r-H m i i 0 f-H CM in 0 O 5- O -C OJ O c— >, ro x -o 0 S- -£Z O CU IE SI C_3 en 10 o IX en 10 CM o f-H CO o 1 X t-H CO »-H O ^_ CM .087C+0. f-H O 0 1 CO UD in cu c cu Q. re X o in 0 iS f-H 0 i-H CO o IX 10 0 I-H

o' CO re re re T? cu •^ f-H I-H ^3 O -r- .x -i CM CM S •-H 0 C? <4- 3 4_> r S- o I-H re 0 $- O J=l 1 re <_> r— CM 1 co -J3 O 3 s cu -C o O O) • -Q O re in o 0 r^ in ex O re CO ai i= ro C o 1 ^ re 1 \ c cu o o •* o in CM CU I-H ._C I | — CO Q- * 505-23


-------

    vt
         ae

         8
         I
     i ••
     lUI
u»s
as
      »-4 t-4 O
  upptaiti
    uppiv
soj^iXdaomo
                       aomoXxoH^K
                      euo^a^t U|jpu3
                                           U|jpu3
                                   Jtomorjng
        jomDE3d*H
 Jom3»Ip5»H


   euepu-ji
    nusqoiHoja
                                                                1
                                                          J.
                                                                1  .
                                                         J
                                                                  J-
                                                              '    1

                                                              *   Iff
                                                                  8J; v» J
                                                                  «2 £^i

                                                              8   § ^5
                                                                          , "o«r
                                                                       z-o
                                                                      *  "5-8 c
   •— c

c«   '"C
                                                                      a
                                    505-24

-------
COLUMN:  Fused silica capillary
LIQUID PHASE:  06-1     F    V
FILM THICKNESS:.  l.Oum
COLUMN DIMENSIONS:  (
                   * * •  •  i
                         10        15      . 20
                                TIKE (KIN)
                                                •  • •
30
35
         Figure  2.   fjtract of reaoent water spiked at 20 ug/L with  atrazlne.
                    60 ug/L with slmazlne, 0.45 ug/L with cls-nonachlor,  and
                    0.35 ug/L with hexachlorocyclopentadlene,  heptachlor,
                    alpha chlprdane, gamma chlordane, and trans-nonachlor.
                           505-25

-------
*
3
       I
       X
       5
       *+
       I
  t~< «-• O
                                         505-26


-------
       I

       R
«»-O  JE
 v>   ••o
   *'' 1'*
U.XOK
   £MQ
 ••   ae
KO^Z
   -ju. 8
                                                                                           &


                                                                                           *

                                                                                           2
                                                                                                  |
                                                                                                  TL
                                                                                                  «t

                                                                                                  S
                                                                                           «      C
                                             505-27

-------
       "
O t-«»-t
                                               505-28

-------
      I
5S53
o •-* *-• o
O^U.0
8


8
                                                                           •)


                                                                           1
                                                                     n
                                505-29

-------
b     ".
<•      o
**v *
 3

 3
*- A »- • •
*r* S   tt
 *"
 ^er^
 O 1-11-«
                                                505-30


-------
      I
      ac

b    ^
2    e

1  :l
»•• I r- ••
  *• TT? fc*
•*  5C


Is   i

OMMO
                                                                          £
                                                                         2    S
                                                                               •0
                                                                               X
                                                                               00
                                  505-31

-------
       I
       ;
VI O   5
     •• fy

*O   »-*
U ••&> tA
MhJUJ Z
s

s

g

*

                                                                                     1

                                                                                                 0*
                                            505-32

-------

 3    fi

 3  I-
*"T *
r~ I r« ••
Tig  g
..

^°  5
gisi
o «-« *-• o
U-IU.O
                                                                   n
                                                                   W ^^
                                                                  n
                                                                  M


                                                                  O


-------
     I
     X

1   I

3  3°
»- » r-
«r" ca
MO
    as
    s
o »-«>-• o
                                  505-34

-------
METHOD 506.  Determination of Phthalate and Adi pate Esters in Drinking Water
             by Liquid-Liquid Extraction or Liquid-Solid Extraction and Gas
             Chromatography with Photoionization Detection
                                 Revision  1.1




                          Edited  by  J.W. Munch  (1995)



       F. K. Kawahara, J. W. Hodgeson - Method 506 Revision  1.0  (1990)
                    NATIONAL EXPOSURE RESEARCH LABORATORY
                     OFFICE  OF  RESEARCH AND DEVELOPMENT
                    U.S.  ENVIRONMENTAL PROTECTION AGENCY
                           CINCINNATI, OHIO 45268
                                    506-1

-------
                              METHOD 506

  Determination  of  Phthalate  and Adipate Esters in Drinking Water
     by Liquid-Liquid Extraction or Liquid-Solid Extraction and
          Gas Chromatography with Photoionization  Detection


SCOPE AND APPLICATION

1.1  This method describes  a  procedure for the determination of certain
     phthalate and  adipate  esters  in drinking water by liquid/liquid or
     liquid/solid extraction.  The following compounds can be determined
     by this method:
1.2
1.3
1.4
               Arialvte

     Bis (2-ethylhexyl) phthalate
     Butyl benzyl phthalate
     Di-n-butyl phthalate
     Diethyl phthalate
     Dimethyl phthalate
     Bis(2-ethylhexyl) adipate
     Di-n-octyl phthalate
                                          Chemical Abstract Services
                                                Registry Number

                                                 117-81-7
                                                 85-68-7
                                                 84-74-2
                                                 84-66-2
                                                 131-11-3
                                                 103-23-1
                                                 117-81-7
     This  is  a capillary column gas chromatographic  (GC) method
     applicable to the determination of the compounds listed above in
     ground water and finished drinking water.  When this method is used
     to analyze unfamiliar  samples'for any or al,l of the compounds listed
     above, compound identifications should be,supported by at least one
     additional qualitative technique.  Method 525.,2 provides gas
     chromatograph/mass spectrometer (GC/MS) conditions appropriate for
     the qualitative and quantitative confirmation of results for all the
     analytes listed above, using the extract produced by this method.

     This method has been validated in a single laboratory, and method
     detection limits (MDLs)  (1) have been determined for the analytes  •
     above (Table 2).  Observed detection limits may vary among waters,
     depending upon the nature of interferences in the sample matrix and
     the specific instrumentation used.

     This method is restricted to use by or under the supervision of
     analysts experienced in  the use of GC, and in the interpretation of
     gas chromatograms obtained by a computerized system.  Each analyst
     must demonstrate the ability to generate acceptable results with
     this method using the procedure described in Sect. 10.

SUMMARY OF METHOD

2.1  A measured volume of sample, approximately 1-L,  is extracted with
     methylene chloride followed by hexane using a glass separatory
     funnel.   The solvent extract is isolated, dried and concentrated to
                                506-2

-------
          a volume of 5 ml or less.  The extract is further concentrated by
          using a gentle stream of nitrogen gas to reduce the sample volume to
          1 ml or less.

          Alternatively, a measured volume of sample is extracted with a
          liquid-solid extraction (LSE) cartridge or disk.  The LSE media are
          eluted with acetonitrile followed by methylene chloride (disk
          extraction) or with methylene chloride only (cartridge extraction)
          The eluant is concentrated using a gentle stream of nitrogen gas or
          clean air to reduce the volume to 1 ml or less.

          The analytes in the extract are separated by means of capillary gas
          chromatography using temperature programming.   The
          chromatographically separated phthalate and adipate esters are
          measured with a photoionization detector, which is operating at 10
3.
DEFINITIONS
     3.1
     3.2
     3.3
     3.4
    3.5
     LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water that
     is treated exactly as a sample including exposure to all  glassware,
     equipment, solvents, reagents, internal  standards,  and surrogates
     that are used with other samples.  The LRB is used  to determine if
     method analytes or other interferences are present  in the laboratory
     environment,  the reagents,  or the apparatus.

     FIELD REAGENT BLANK (FRB), — Reagent water placed in a sample
     container in  the laboratory and treated  as a sample in all  respects,
     including exposure to sampling site conditions,  storage,
     preservation  and all analytical procedures.  The  purpose of the FRB
     is to determine if method analytes or other  interferences are
     present in the field environment.

     LABORATORY FORTIFIED BLANK  (LFB)  - An aliquot of reagent water to
     which known quantities of the method analytes are added in  the
     laboratory.   The LFB is analyzed  exactly Tike a  sample,  and its
     purpose is to determine whether the methodology  is  in control,  and
     whether the laboratory is capable of making  accurate and  precise
     measurements  at the required method detection limit.

     LABORATORY FORTIFIED SAMPLE MATRIX (LFM)  —  An aliquot of an
     environmental  sample to which known quantities of the method
     analytes  are  added in the laboratory.  The LFM is analyzed  exactly
     like  a  sample,  and its purpose is to determine whether the  sample
     matrix  contributes bias to  the analytical  results.  The background
     concentrations  of the analytes in the  sample  matrix  must  be
     determined in  a separate aliquot  and the  measured values  in the LFM
     corrected  for  background concentrations.

     STOCK STANDARD  SOLUTION --  A concentrated solution  containing  a
     single  certified  standard that is  a method analyte,  or a
     concentrated  solution  of a  single  analyte prepared  in  the laboratory

                               506-3

-------
          with an assayed reference compound.  Stock standard solutions are
          used to prepare primary dilution standards.

     3.6  PRIMARY DILUTION STANDARD SOLUTION — A solution of several  analytes
          prepared in the laboratory from stock standard solutions and diluted
          as needed to prepare calibration solutions and other needed analyte
          solutions.

     3.7  CALIBRATION STANDARD (CAL) — a, solution prepared from the primary
          dilution standard solution and stock standard solutions of the
          internal standards and surrogate analytes.  The CAL solutions are
          used to calibrate the instrument response with respect to analyte
          concentration.

     3.8  QUALITY CONTROL SAMPLE (QCS) — a sample matrix containing method
          analytes or a solution of method analytes in a water miscible
          solvent which is used to fortify reagent water or environmental
          samples.  The QCS is obtained from a source external to the
          laboratory, and is used to check laboratory performance with
          externally prepared test materials.

4.   INTERFERENCES

     4.1  Method interferences may be caused by contaminants in water,
          solvents, reagents, glassware, and sample processing hardware.
          These lead to discrete artifacts and/or elevated baselines in gas
          chromatograms.  All of these materials must be routinely
          demonstrated to be free from interferences under the conditions of
          the analysis by running laboratory reagent blanks (Sect. 10.2).

          4.1.1  Phthalate esters are contaminants in many products found in
                 the laboratory.  It is particularly important to avoid the
                 use of plastics because phthalates are commonly used as
                 plasticizers and are easily extracted from plastic materials.
                 Great care must be exercised to prevent contamination.
                 Exhaustive clean up of reagents and glassware must be
                 required to eliminate background phthalate that is not
                 derived from the sample.

          4.1.2  Glassware must be scrupulously cleaned.  Clean all glassware
                 as soon as possible after use by thoroughly rinsing with the
                 last solvent used.  Follow by washing with hot water and
                 detergent and thorough rinsing with tap and reagent water.
                 Drain dry and heat in an oven or muffle furnace at 400°C for
                 1 hour.  Do not heat volumetric glassware.  Thorough rinsing
                 with acetone may be substituted for the heating.  After
                 cooling, the glassware should be sealed with aluminum foil
                 and stored in a clean environment to prevent accumulation of
                 dust and other contaminants.
                                     506-4

-------
     4.1.3   The  use  of  high  purity  reagents  and  solvents  helps to
             minimize interference problems.   Purification of  solvents by
             distillation  in  an  all  glass  system  may be required.
             WARNING:  When a solvent  is purified,  stabilizers added by
             the  manufacturer are removed  thus potentially making the
             solvent  hazardous.  Also, when a  solvent  is purified,
             preservatives added by  the manufacturer are removed thus
             potentially reducing the  shelf-life.

4.2  Matrix  interferences may be caused by contaminants that  are co-
     extracted from  the sample.  The  extent of matrix interferences will
     vary from source to  source, dependent upon  the nature and diversity
     of the  samples.  Clean  up  procedures can be used to overcome many of
     these interferences.

4.3  It is important that samples and working standards be contained in
     the same solvent.  The  solvent for working  standards must be the
     same as the final  solvent  used in sample preparation.  If this is
     not the case, chromatographic  comparability of standards to sample
     may be  affected.

SAFETY

5.1  The toxicity or carcinogenicity  of each reagent used in this method
     has not been precisely  defined;  however, each chemical compound must
     be treated  as a potential  health hazard.  Accordingly, exposure to
     these chemicals must be  reduced  to the lowest possible level.  The
     laboratory  is responsible  for maintaining a current awareness file
     of OSHA regulations regarding the safe handling of the chemicals
     specified in this  method.  A reference file of material safety data
     sheets  should also be made available to all personnel involved in
     the chemical analysis.   Additional references to laboratory safety
     are available and  have  been identified (5-7) for the information of
     the analyst.

EQUIPMENT AND SUPPLIES  (All   specifications are suggested, catalog
numbers are  included for illustration only.)

6.1  Sampling Equipment

   .  6.1.1  Grab Sample Bottle—1-L or 1-qt amber glass,  fitted with a
            screw cap lined with Teflon.   Protect samples from light if
            amber bottles are not available.  The bottle and cap liner
            must be washed, rinsed with acetone or methylene chloride and
            dried before use  in order to minimize contamination.   (See
    '•     ,  4.1.1.)

6.2  Glassware

     6.2.1  Separatory  Funnel—2-L with Teflon stopcock.
                                506-5

-------
     6.2.2  Drying Column—Chromatographic column-300 mm long x 10 mm ID,
            with Teflon stopcock and coarse frit filter disc at bottom.

     6.2.3  Concentrator Tube—Kuderna-Danish, 10 ml, graduated,
            calibration must be checked at the volumes employed in the
            test.  Tight ground glass stopper is used to prevent
            evaporation of extracts.

     6.2.4  Evaporative Flask—Kuderna-Danish, 500 ml, attach to
            concentrator tube with springs.

     6.2.5  Snyder Column—Kuderna-Danish, three-ball macro size.

     6.2.6  Snyder Column—Kuderna-Danish, 2 or 3 ball micro size.

     6.2.7  Vials—10 to 15 ml, amber glass with Teflon-lined screw cap.

     6.2.8  Boiling Chips—Approximately 10/40 mesh.   Heat to 400°C for
            30 min. or extract with methylene chloride in a Soxhlet
            apparatus.

     6.2.9  Flask, Erlenmeyer—250 ml.

     6.2.10 Chromatography column similar to 6.2.2.

     6.2.11 Pasteur Pipets (and Bulb).

     6.2.12 Autosampler Vials—Equipped with Teflon-lined septum  and
            threaded or crimp top caps.

6.3  Water Bath—Heated (with concentric ring covers) capable of
     temperature control (± 2°C).  The water bath should be used  in a
     ventilating hood.

6.4  Balance—Analytical,  capable of weighing accurately to nearest
     0.0001 gm.

6.5  Gas Chromatograph—An analytical system complete with temperature
     programmable GC fitted with split-splitless injection mode system,
     suitable for use with capillary columns and all  required accessory
     syringes, analytical  columns, gases, detector and stripchart
     recorder.  A data system for processing chromatographic data  is
     recommended.

     6.5.1  Column, Fused Silica Capillary—DB-5 or equivalent, 30 m long
            x 0.32 mm ID with a film thickness of 0.25 micron.

     6.5.2  The alternate column, Fused Silica Capillary—30 m long x
            0.32 mm ID with a film thickness of 0.25  micron, DB-1  or
            equivalent.
                                506-6

-------
          6.5.3  Detector— A high temperature photoionization detector
                 equipped for 10 electron volts (nominal voltage) and capable
                 of operating from 250°C to 350°C is required.

          6.5.4 'An automatic injector system is suggested, but was not used
                 for the development of this method.

     6.6  Vacuum source, capable of maintaining a vacuum of 10-15 mm Hg.

7.   REAGENTS AND STANDARDS

     7.1  Reagent Water -- Reagent water is defined as water in which an
          interfering substance is not observed at the MDL of the parameters
          of interest.  Reagent water used to generate data in this method was
       .   distilled water obtained from the Millipore L/A-7044 system
          comprised of prefiltration, organic adsorption, deionization and
          Millipore filtration columnar units. Any system may be used if it
          generates acceptable reagent water.

     7.2  Acetone, hexane, methylene chloride, ethyl acetate, ethyl ether and
          iso-octane — Pesticide quality or equivalent to distillation in
          glass quality.

     7.3  Sodium Sulfate—(ACS) Granular, anhydrous.  Several levels of
          purification may be required in order to reduce background phthalate
          levels towards acceptance:  1) Heat 4 h at 400°C in a shallow tray,
          2) Soxhlet extract with methylene chloride for 48 h.

     7.4  Florisil—PR grade (60/100 mesh).  To prepare for use, place 100 g
          of Florisil  into a 500-mL beaker and heat for approximately 16 h at
          40°C.  After heating transfer to a 500-mL reagent bottle.  Tightly
          seal-and cool to room temperature.  When1cool, add 3 ml of reagent
          water.  Mix thoroughly by shaking or rolling for 10 min.  and let it
          stand for at least 2 h.  Store in the dark in glass containers with
          ground glass stoppers or foil-lined screw caps.

     7.5  Sodium Chloride—(ACS) Granular.   Heat 4 h at 400°C in a  shallow
          tray.  When cool,  keep in tightly sealed glass (not plastic) bottle.
          This  cleaning step is required to .minimize background contamination
          associated with this reagent.

     7.6  Ethyl Ether—(ACS) reagent grade.

     7.7  Sodium Thiosulfate (Na2S203) —(ACS) reagent grade.

     7.8  Alumina—Neutral activity Super I, W200 series (ICN Life  Sciences
          Group, No. 404583).  To prepare for use,  place 100 g of alumina into
          a 500-mL beaker and heat for approximately 16 h at 400°C.  After
          heating transfer to a 500-mL reagent bottle.   Tightly seal and cool
          to room temperature.   When cool,  add 3 mL of reagent water.   Mix
          thoroughly by shaking or rolling  for 10 min.  and let it stand for at
          least 2 h.  Keep the bottle sealed tightly.

                                     506-7

-------
     7.9   Liquid-solid  extraction  (LSE)  cartridges.  Cartridges are inert non-
           leaching  plastic,  for  example  polypropylene, or glass, and must not
           contain plasticizers,  such  as  phthalate esters or adipates, that
           leach  into methylene chloride.  The cartridges are packed with about
           1 gram of silica,  or other  inert  inorganic support, whose surface is
           modified  by chemically bonded  octadecyl (C18)  groups.   The  packing
           must have a narrow size  distribution and must not leach organic
           compounds into methylene chloride.  One liter of water should pass
           through the cartridge  in about 2  hrs with the assistance of a slight
           vacuum of about  13 cm  (5 in.)  of  mercury.  The extraction time
           should not vary  unreasonably among LSE cartridges.

     7.10  Liquid-solid  extraction  disks, C-18, 47 mm.  Disks are manufactured
           with Teflon or other inert  support and should contain very little
           contamination.

     7.11  Helium carrier gas, as contaminant free as possible.

     7.12  Stock  standard solutions (1.00 /^g/^L)  - Stock  standard  solutions
           can be prepared  from pure standard materials or purchased as
           certified solutions.

           7.12.1 Prepare stock standard  solutions by accurately weighing
                 about  0.0100 g  of pure material.  Dissolve the material in
                 isooctane and dilute to volume in a 10-mL volumetric flask.
                 Larger volumes  can be used at the convenience of the analyst.
                 When compound purity is assayed to be 96% or greater, the
                 weight can  be used without correction to calculate the
                 concentration of  the stock standard.  Commercially prepared
                 stock  standards can be used at any concentration if they are
                 certified by the  manufacturer or by an independent source.

           7.12.2 Transfer  the stock standard solutions into Teflon-sealed
                 screw-cap bottles.  Store  at 4°C and protect from light.
                 Stock  standard  solutions should be checked frequently for
                 signs  of degradation or evaporation, especially just prior to
                 preparing calibration standards from them.

           7.12.3 Stock  standard  solutions must be replaced after six months,
                 or sooner if comparison with check standards indicates a
                 problem.  Butyl benzyl phthalate is especially vulnerable to
                 autoxidation.

8.   SAMPLE COLLECTION. PRESERVATION. AND STORAGE

     8.1  Grab samples  must  be collected in amber glass containers (Sect.6.1).
          Conventional   sampling  practices should be followed (8,9);  however,
          the bottle must not be prerinsed with sample before collection.

     8.2  SAMPLE PRESERVATION AND  STORAGE
                                     506-8

-------
          8.2.1  For sample dechlorination, add 80 mg sodium thiosulfate to
                 the sample bottle at the sampling site or in the laboratory
                 before shipping to the sampling site.

          8.2.2  After the sample is collected, seal the bottle and swirl the
                 sample until the thiosulfate is dissolved.

          8.2.3  The samples must be iced or refrigerated at 4°C free from
                 light from the time of collection until extraction.  Limited
                 holding studies have indicated that the analytes thus stored
                 are stable up to 14 days or longer. ,Analyte stability may be
                 affected by the matrix; therefore1, the analyst should verify
                 that the preservation technique is applicable to the
                 particular samples under study.

     8.3  Extract Storage — Extracts should be stored at 4°C in absence of
          light. A 14-day maximum extract storage time is recommended.  The
          analyst should verify appropriate extract holding times applicable
          to the samples under study.

9.   QUALITY CONTROL

     9.1  Minimum quality control  (QC) requirements are initial  demonstration
          of laboratory capability,  analysis of laboratory reagent blanks,
          laboratory fortified samples,  laboratory fortified blanks, and QC
          samples.   A MDL for each analyte must also be determined.
          Additional  quality control  practices are recommended.

     9.2  Laboratory Reagent Blanks (LRB)   Before processing any samples, the
          analyst must demonstrate that all  glassware and reagent
          interferences are under control.  Each time a set of samples is
          extracted or reagents are changed, a LRB must be analyzed.  If
          within the retention time window of any analyte of interest the LRB
          produces  a peak that would prevent the determination of that analyte
          using a known standard,  determine the source of contamination and
          eliminate the interference before processing samples.

     9.3  Initial Demonstration of Capability.

          9.3.1  Select a representative fortified concentration (about 10
                 times EDL or at a concentration in the middle of the
                 calibration range established in Section 10)  for each
                 analyte.   Prepare a primary dilution standard (in methanol)
                 containing each analyte at  1000 times selected  concentration.
                 With a syringe,  add  1  mL  of the concentrate to  each of four
                 to seven 1-L aliquots  of  reagent water,  and analyze each
                 aliquot according to procedures beginning in  Sect.  11.

          9.3.2  For  each analyte  the mean recovery value for  these  samples
                 must fall  in the  range  of R ± 30% using  the values  for R for
                 reagent water in  Tables 3 or 4.   The precision  of these
                 measurements,  expressed as  RSD,  must be  20% or  less.   For

                                    506-9

-------
            those compounds that meet the acceptance criteria,
            performance is considered acceptable.   For those compounds
            that fail  these criteria, this procedure must be repeated
            using fresh replicate samples until  satisfactory performance
            has been demonstrated.

     9.3.3  For each analyte,  determine the MDL.   Prepare a minimum of  7
            LFBs at a low concentration.   Fortification concentration in
            Table 2 may be used as  a guide, or use calibration  data
            obtained in Section 10  to estimate a concentration  for each
            analyte that will  produce a peak with a 3-5 times signal  to
            noise response.  Extract and analyze each replicate according
            to Sections 11 and 12.   It is recommended that these LFBs be
            prepared and analyzed over a period of several days, so that
            day to day variations are reflected in precision
            measurements.  Calculate mean recovery and standard deviation
            for each analyte.  Use the standard deviation and the equation
            given in Section 13 to  calculate the MDL.

     9.3.4  The initial demonstration of capability is used primarily to
            preclude a laboratory from analyzing unknown samples via a
            new, unfamiliar method  prior to obtaining some experience
            with it.  It is expected that as laboratory personnel gain
            experience with this method the quality of data will improve
            beyond those required here.

9.4  The analyst is permitted to modify GC columns, GC conditions,
     concentration techniques (i.e. evaporation techniques), internal
     standards or surrogate compounds.  Each time such method
     modifications are made, the analyst must repeat the procedures in
     Sect. 9.3.

9.5  Assessing Laboratory  Performance - Laboratory Fortified Blank

     9.5.1  The laboratory must analyze at least one laboratory fortified
            blank (LFB) sample with every twenty samples or one per
            sample set (all samples extracted within a 24-h period)
            whichever is greater.  Ideally, the fortified concentration
            of each analyte in the LFB should be the same concentration
            selected in Section 9.3.1.  Calculate accuracy as percent
            recovery (X-)-  If the  recovery of any analyte falls outside
            the control limits  (see Sect. 9.5.2), that analyte  is judged
            out of control, and the  source of the problem should be
            identified and resolved  before continuing analyses.

     9.5.2  Until sufficient data become  available from within  their own
            laboratory, usually a minimum of results from 20 to 30
            analyses, the  laboratory  should assess laboratory performance
            against the control limits in Sect. 9,3.2 that are  derived
            from the data  in Table 2.  When sufficient internal
            performance data becomes  available, develop  control limits
            from the mean  percent recovery  (X) and standard deviation  (S)

                                506-10

-------
                  of the percent recovery.   These  data  are  used  to  establish
                  upper and lower control  limits as  follows:

                            UPPER CONTROL  LIMIT = X +  3S               *
                            LOWER CONTROL  LIMIT =  X -  3S

                  After each five to  ten new recovery measurements,  new control
                  limits should  be calculated using  only the most recent  20-30
                  data  points.   These calculated control limits  should not
                  exceed those established  in Sect.  9.3.2.

     9.6  Assessing Analyte Recovery — Laboratory  Fortified  Sample Matrix

          9.6.1   The laboratory must fortify each analyte  to  a  minimum of 10%
                  of the routine samples or  one fortified sample per set,
                  whichever is greater.  The fortified  concentration should not
                  be less  than the background concentration of the  sample
                  selected for fortifying.   Ideally,  this concentration should
                  be the same as that used for the laboratory  fortified blank
                  (Sect.  9.5).   Over  time, samples from all routine  sample
                  sources  should be fortified.

          9.6.2   Calculate the  accuracy as  percent  recovery,  R, for each
                  analyte,  corrected  for background  concentrations measured in
                  the unfortified  sample.  For each  analyte the  mean recovery
                  value  for these  samples must fall  in  the range of  R ± 35%
                  using  the values  for R for reagent water in  Tables 3 or 4.

          9.6.3   If the recovery  of  any such  analyte falls outside  the
                  designated range, and the  laboratory  performance for that
                  analyte  is shown to be in  control  (Sect. 9.5), the recovery
                  problem  encountered with the  dosed sample is judged to be
                  matrix related,  not system related.   The result for that
                  analyte  in the unfortified  sample  is  labeled suspect/matrix
                  to  inform the  data  user that  the results are suspect due to
                  matrix effects.

          QUALITY CONTROL  SAMPLES  (QCS) - Each quarter, the laboratory should
          analyze one or more QCS  (if available).    If  criteria provided with
          the QCS are not  met,   corrective action  should be taken and
          documented.

          The laboratory may adopt additional quality control  practices for
          use with this method.   The specific practices that are most
          productive  depend upon the needs of the laboratory and the nature
          of the  samples.  For example,  field or laboratory duplicates may be
          analyzed to assess the precision of the  environmental  measurements.

10.   CALIBRATION AND STANDARDIZATION

     10.1  Establish gas chromatograph operating conditions  equivalent to those
          given in Table 1.  The gas chromatographic system is calibrated

                                    506-11
9.7
9.8

-------
     using the external standard technique.   Calibration standards must
     be prepared in the same solvent as the  final  sample extract.   This
     will vary with the extraction option chosen (hexane for LLE,
     methylene chloride for LSE-cartridge,  and acetonitrile for LSE-
     disk).  Sect. 10.2.1 details the hexane option as an example.

10.2 External standard calibration procedure:

     10.2.1 Prepare calibration standards at a minimum of three
            concentration levels for each analyte of interest by adding
            volumes of one or more stock standards to a volumetric flask
            and diluting to volume with n-hexane.   Guidance on the number
            of standards is as follows:  A minimum of three calibration
            standards are required to calibrate a range of a factor of 20
            in concentration.  For a factor of 50 use at least four
            standards, and for a factor of 100 at least five standards.
            One calibration standard should contain each analyte of
            concern at a concentration 2 to 10 times greater than the
            method detection limit for that compound.  The other
            calibration standards should contain each analyte of concern
            at concentrations that define the range of the sample analyte
            concentrations or should define the working range of the
            detector.

     10.2.2 Using injections of 1 to 2 jLtL,  analyze each calibration
            standard according to Sect. 11.5 and tabulate peak height or
            area responses against the mass injected.  The results can be
            used to prepare a calibration curve for each compound.
            Alternatively, if the ratio of response to amount injected
            (calibration factor) is a constant over the working range
            (<20% relative standard deviation, RSD), linearity through
            the origin can be assumed and the average ratio or
            calibration factor can be used in place of a calibration
            curve.

     10.2.3 The working calibration curve or calibration factor must be
            verified on each working day by the measurement of a minimum
            of two calibration check standards, one at the beginning and
            one at the end of the analysis day.  These check standards
            should be at two different concentration levels to verify the
            calibration curve.  For extended periods of analysis  (greater
            than 8 hrs.), it is strongly recommended that check standards
            be interspersed with samples at regular intervals during the
            course of the analyses.  If the response for any analyte
            varies from the predicted response by more than ±20%, the
            test must be repeated using a fresh calibration standard.  If
            the results still do not agree, generate a new calibration
            curve.  For those analytes that failed the calibration
            verification, results from field samples analyzed since the
            last passing calibration should be considered suspect.
            Reanalyze sample extracts for these analytes after acceptable
            calibration is restored.

                               506-12

-------
11.   PROCEDURE

     11.1 LIQUID-LIQUID EXTRACTION

          11.1.1 Mark the water meniscus on the side of the sample bottle for
                 later determination of sample volume.   Pour the entire sample
                 into a 2-L separatory funnel  containing 50 g of NaCl.

          11.1.2 Add 60 ml  CH2C12( to the sample bottle.  Seal, and shake
                .gently to rinse'the inner walls of the bottle.  Transfer the
                 solvent to the separatory funnel.   Extract the  sample  by
                 shaking the funnel  for 2 min  with  initial  and periodic
                 venting to release  excess pressure.  Allow the  organic layer
                 to separate for a minimum of  10 min from the water phase.   If
                 the emulsion interface between layers  is more than one-third
                 the volume of the solvent layer,  the analyst must employ
                 mechanical techniques to complete  the  phase separation.  The
                 optimum technique depends upon the sample, but  may include
                 stirring,  filtration of the emulsion through glass wool,
                 centrifugation, or  other physical  methods.  Collect the
                 solvent extract in  a 250-mL Erlenmeyer flask.

          11.1.3 Add a second 60-mL  volume of  methylene chloride to the sample
                 bottle and repeat the extraction  procedure a second time,-
                 combining the extracts in the Erlenmeyer flask.   Perform a
                 third extraction in the same  manner.  Then extract with 40-mL
                 of hexane, which extract (top phase) is added to the total.

          11.1.4 Assemble a Kuderna-Danish (K-D) concentrator by attaching  a
                 10-mL concentrator  tube to a  500-mL evaporative flask.   Other
                 concentration devices or techniques may be used in place of
                 the K-D concentrator,  provided the concentration factor
                 attained in 11.1.6  - 11.1.8 is achieved without loss of
                 analytes.

          11.1.5 Pour the combined extract through  a drying column (6.2.2)
                 containing about 10 cm of prerinsed .anhydrous sodium sulfate,
                 and collect the extract in the K-D concentrator.   Rinse the
                 Erlenmeyer flask and column with 20 to 30  mL of methylene
                 chloride to complete the quantitative  transfer.

          11.1.6 Add one or two clean boiling  chips to  the  evaporative  flask
                 and attach a three-ball  Snyder column.   Prewet  the Snyder
                 column by  adding about 1  mL of methylene chloride to the top.
                 Place the  K-D apparatus on a  hot water bath (60  to 65°C) so
                 that the concentrator tube is partially immersed in the hot
                 water,  and the entire lower rounded surface of  the flask  is
                 bathed with hot vapor.   Adjust the vertical  position of the
                 apparatus  and the water temperature as  required  to complete
                 the concentration in 40 min.   At the proper rate of
                 distillation the balls of the column will  actively chatter
                 but the chambers will  not flood with condensed  solvent.  When

                                    506-13

-------
            the apparent volume of liquid reaches approximately 7 ml,
            remove the K-D apparatus and allow it to drain and cool  for
            at least 10 min.

     11.1.7 Increase the temperature of the hot water bath to about  85°C.
            Remove the Snyder column, rinse the column and the 500-mL
            evaporative flask with 1 - 2 ml of hexane.  Replace with a
            micro column and evaporative flask.  Concentrate the extract
            as in Sect. 11.1.6 to 0.5 - 1 ml_.  The elapsed time of
            concentration should be approximately 15 min.

     11.1.8 Remove the micro Snyder column and rinse the column by
            flushing with hexane using a 5-mL syringe.  Concentrate  to a
            volume of 1 ml by purging the liquid surface with a gentle
            flow of nitrogen or clean air. If an autosampler is to be
            used, transfer the extract to an autosampler vial with a
            Pasteur pipet.  Seal the vial with a threaded or crimp top
            cap.  Store in refrigerator if further processing will not be
            performed.  If the sample extract requires no further
            cleanup, proceed with gas chromatographic analysis (Sect.
            11.5).  If the sample requires further cleanup, proceed  to
            Sect. 11.4.

     11.1.9 Determine the original sample volume by refilling the sample
            bottle to the mark and transferring the liquid to a 1000-mL
            graduated cylinder.  Record the sample volume to the nearest
            5 ml.

11.2 LIQUID-SOLID EXTRACTION - CARTRIDGE OPTION

     11.2.1 This method is applicable to a wide range of organic
            compounds that are efficiently partitioned from the water
            sample onto a C18  organic phase  chemically bonded  to  a solid
            inorganic matrix, and are sufficiently volatile and thermally
            stable for gas chromatography (10).  See Section 11.3 for the
            disk option procedure.  Particulate bound organic matter will
            not be partitioned, and more than trace levels of
            particulates in the water may disrupt the partitioning
            process.  Single laboratory accuracy and precision data have
            been determined at a single concentration for the analytes
            listed in Sect. 1.1 fortified into reagent water and raw
            source water.

     11.2.2 Set up the extraction apparatus shown in Figure 1A. An
            automated extraction system may also be used. The reservoir
            is not required, but recommended for convenient operation.
            Water drains from the reservoir through the LSE cartridge and
            into a syringe needle which is inserted through a rubber
            stopper into the suction flask.  A slight vacuum of 13 cm (5
            in.) of mercury is used during all operations with the
            apparatus.  With this extraction apparatus, sample elution
            requires approximately 2 hours.  Acceptable new cartridge and

                               506-14

-------
       extraction, disk technology have recently become available,
       which allow significantly faster elution rates.

11.2.3 Mark the water meniscus on the side of the sample bottle
       (approximately 1 liter) for later determination of sample
       volume.  Pour the water sample into the 2-1 separatory funnel
       with the stopcock closed.

11.2.4 Flush each cartridge with two 10 ml aliquots of methylene
       chloride, followed by two 10 mL aliquots of methanol, letting
       the cartridge drain dry after each flush.  These solvent
       flushes may be accomplished by adding the solvents directly
       to the solvent reservoir in Figure 1A.  Add 10-mL of reagent
       water to the solvent reservoir, but before the reagent water
       level drops below the top edge of the packing in the LSE
       cartridge,  open the stopcock of the separatory funnel and
    ,   begin adding sample water to the solvent reservoir.   Close
       the stopcock when an adequate amount of sample is in the
       reservoir.                                   .    '

11.2.5 Periodically open the stopcock and drain a portion of the
       sample water into the solvent reservoir.  The water  sample
       will  drain  into the cartridge, and from the exit into the
       suction flask.   Maintain the packing material  in the
       cartridge immersed in water at all  times.  After all of the
       sample has  passed .through the LSE cartridge,  wash the
       separatory  funnel  and cartridge with 10 ml of reagent water,
       and draw air through the cartridge for about 10 min.

11.2.6 Transfer the 125-mL solvent reservoir and LSE cartridge (from
       Figure 1A)  to the elution apparatus (Figure IB).   The same
       125 mL solvent  reservoir is used for both apparatus.  Wash
       the 2-liter separatory funnel  with  5 mL of methylene chloride
       and collect the washings.   Close the stopcock on the 100-mL
       separatory  funnel  of the elution apparatus,  add the  washings
       to the reservoir and enough additional  methylene chloride to
       bring the volume back up to 5 mL and elute the LSE cartridge.
       Elute the LSE cartridge with an additional  5  mL of methylene
       chloride (10-mL total).   A small  amount of nitrogen  positive
       pressure may be used to elute the cartridge.  Small amounts of
       residual  water  from the LSE cartridge will  form an immiscible
       layer with  the  methylene chloride in the 100-mL separatory
       funnel.   Open the  stopcock and allow the methylene chloride
       to pass  through the drying column packed with  anhydrous
       sodium sulfate  (1-in)  and  into the  collection  vial.   Do not
       allow the water layer to enter the  drying column.  Remove the
       100 mL separatory  funnel  and wash the drying  column  with  2 mL
       of methylene chloride.   Add this  to the extract.   Concentrate
       the extract  to  1  mL under  a gentle  stream:of  nitrogen.  The
       extract  is  now  ready for gas chrbmatqgraphy  (Sect. 11.4)  or
       additional  cleanup  (Sect.  11.3).
                          506-15

-------
11.3 LIQUID-SOLID EXTRACTION - DISK OPTION

     11.3.1 Preparation of disks.

            11.3.1.1 Insert the disk into the 47 mm filter apparatus.
                     Wash the disk with 5 mL methylene chloride (MeCl2)
                     by adding the MeC12 to the disk,  pulling  about  half
                     through the disk and allowing it to soak the disk
                     for about a minute, then pulling the remaining  MeCl2
                     through the disk.  With the vacuum on,  pull  air
                     through the disk for a minute.

            11.3.1.2 Pre-wet the disk with 5 mL methanol (MeOH) by adding
                     the MeOH to the disk, pulling about half through  the
                     disk and allowing it to soak for about  a minute,
                     then pulling most of the remaining MeOH through.   A
                     layer of MeOH must be left on the surface of the
                     disk, which shouldn't be allowed to go  dry from this
                     point until the end of the sample extraction.  THIS
                     IS A CRITICAL STEP FOR A UNIFORM FLOW AND GOOD
                     RECOVERY.

            11.3.1.3 Rinse the disk with 5 mL reagent water  by adding  the
                     water to the disk and pulling most through,  again
                     leaving a layer on the surface of the disk.

     11.3.2 Add 5 mL MeOH per liter of water sample.   Mix well.

     11.3.3 Add the water sample to the reservoir'and turn 'on the vacuum
            to begin the filtration.  Full aspirator vacuum  may be used.
            Particulate-free water may filter in as little as 10  minutes
            or less.  Filter the entire sample, draining as  much  water
            from the sample container as possible.

     11.3.4 Remove the filtration top from the vacuum flask, but  don't
            disassemble the reservoir and fritted base.  Empty the water
            from the flask and insert a suitable sample tube to contain
            the eluant.  The only constraint on the sample tube is that
            it fit around the drip tip of the fritted base.   Reassemble
            the apparatus.

            Add 5 mL of acetonitrile (CH3CN)  to rinse  the  sample  bottle.
            Allow the CH,CN  to settle to the  bottom of the bottle  and
            transfer to the disk with a dispo-pipet,  rinsing the  sides of
            the glass filtration reservoir in the process.   Pull  about
            half of the CH3CN through the disk,  release the  vacuum,  and
            allow the disk to soak for a minute.  Pull the remaining
            CH3CN through  the disk.

            Repeat the above step twice, using MeCl,  instead of CH3CN.
            Pour the combined eluates thru a small funnel  with filter
            paper containing 3 grams of anhydrous sodium sulfate.  Rinse

                               506-16

-------
            the test tube and sodium sulfate with two 5 mL portions of
            MeCl2.   Collect  the  filtrate in a concentrator tube.

     11.3.5 With the concentrator tube in a 28°C heating block,
            evaporate the eluate with a stream of N2  to 0.5 ml.'

11.4 EXTRACT CLEANUP - Cleanup procedures may not be  necessary for a
     relatively clean sample matrix,  such as most drinking waters.  If
     particular circumstances demand  the use of a  cleanup procedure,  the
     analyst may use either  procedure below or any other appropriate
     procedure.  However, the analyst first must  demonstrate that the
     requirements of Sect. 9 can be met using the method as revised to
     incorporate the cleanup procedure.

     11.4.1 Florisil column  cleanup for phthalate esters:

            11.4.1.1 Place 10 g  of Florisil  (see 7.4)  into a
                     chromatographic  column.   Tap the  column to settle
                     the Florisil  and add 1  cm of anhydrous sodium
                     sulfate to  the top.

            11.4.1.2 Preelute the  column  with 40 mL of hexane.   Discard
                     the eluate  and just  prior to exposure  of the  sodium
                     sulfate  layer to  the  air,  quantitatively  transfer
                     the sample  extract (11.1.8.or 11.2.6)  onto the
                     column,  using  an additional  2 mL  of hexane to
                     complete the  transfer.   Just prior to  exposure  of
                     the sodium  sulfate layer to the air,  add 40 mL  of
                     hexane  and  continue  the  elution of the  column.
                     Discard  this  hexane  eluate.

            11.4.1.3  Next, elute the  column with  100 mL  of  20%  ethyl
                     ether in  hexane (V/V) into  a 500-mL K-D flask
                     equipped with  a  10-mL concentrator  tube.   Elute the
                     column  at a rate of  about 2  mL/min  for  all
                     fractions.  Concentrate  the  collected  fraction  as in
                     Section  11.1.  No  solvent exchange  is necessary.
                     Adjust the volume  of the cleaned extract to 1 mL  in
                     the  concentrator tube and analyze by gas
                     chromatography.

    11.4.2 Alumina column cleanup  for  phthalate  esters:

           11.4.2.1  Place 10 g of  alumina into a chromatographic column-
                    Tap the  column to  settle the alumina and add 1 cm of
                     anhydrous sodium sulfate to the top.

           11.4.2.2 Preelute the column with 40 mL of hexane.  The rate
                    for all  elutions  should be about  2 mL/min.  Discard
                    the eluate and just prior to exposure of the sodium
                    sulfate  layer to  the air, quantitatively transfer
                              506-17

-------
                     the sample extract (Sect. 11.1.8 or 11.2.6)  onto the
                     column, using an additional  2 mL of hexane to
                     complete the transfer.  Just prior to exposure of
                     the sodium sulfate layer to  the air, add 35  ml of
                     hexane and continue the elution of the column.
                     Discard this.hexane eluate.

            11.4.2.3 Next, elute the column with  140 mL of 20% ethyl
                     ether in hexane (V/V) intq a 500-mL K-D flask
                     equipped with a 10-mL concentrator tube. Concentrate
                     the collected fraction as in Section 11.1.  No
                     solvent exchange is necessary.  Adjust the volume of
                     the cleaned extract to 1 ml  in the concentrator tube
                     and analyze by gas chromatography.

11.5 GAS CHROMATOGRAPHY

     11.5.1 Table 1 summarizes the recommended operating conditions for
            the gas chrpmatograph.  Included are  retention data for the
            primary and confirmation columns.  Other capillary columns,
            chromatographic conditions may be used if the requirements of
            Section 9 are met.

     11.5.2 Calibrate the system daily as described in Sect. 10.'

     11.5.3 Inject 1 to 2 /iL of the sample extract or standard into the
            gas chromatograph.  Smaller (1.0 /til)  volumes may be injected
            if automatic devices are employed.  For optimum
            reproducibility, an autoinjector is recommended.

     11.5.4 Identify the analytes in the sample by comparing the
            retention times of the peaks in the sample chromatogram with
            those of the peaks in standard chromatograms.  The width of
            the retention time window used to make identifications should
            be based upon measurements of actual  retention time
            variations of standards over the course of a day.  Three
            times the standard deviation of a retention time for  a
            compound can be used to calculate a suggested window  size;
            however, the experience of the analyst should weigh heavily
            in the interpretation of chromatograms.

     11.5.5 If the response ,for a peak exceeds the working range  of the
            system, dilute the extract and reanalyze.

     11.5.6 If the measurement of the peak response is prevented  by the
            presence of interferences, further cleanup is required.

     11.5.7 The calibration curves should be linear over the range of
            concentrations in Tables 2-5.   Do not extrapolate beyond the
            calibration range established in Section 10.  If analyte
            response is too high, dilute the extract and reanalyze.
                               506-18

-------
12.   DATA ANALYSIS AND CALCULATIONS

     12.1 Calculate the amount of material injected from the peak response
          using the multi-point calibration curve or calibration factor
          determined in Section 10.2.2.  Do "not use the daily calibration
          verification standard to quantitate method analytes in samples.
          concentration in the sample can be calculated from Equation 2.
                                                                            The
                Equation 2.
                  where:
                               Concentration  (Mg/L) =  4§*£U_
                    A  = Amount of material  injected (ng).
                    V,.  = Volume of extract  injected  (fj.1).
                    Vt  = Volume of total  extract  (jiL).
                    Vs  = Volume of water  extracted  (mL).

      12.2  Report  results in Aig/L without correction for recovery data,
           QC  data obtained should be reported with  the sample results.

 13.   METHOD PERFORMANCE
                                                                        All
      13.1  Single  laboratory accuracy and precision data were obtained by
           replicate  liquid-liquid extraction  analyses of reagent water
           fortified  at  two  sets  of concentrations  of method analytes.  The
           data  are given  in Tables 2 and 3.   Accuracy and precision data by
           liquid-solid  extraction of reagent  water fortified at a single
           concentration are given in Table  4.   Finally,  Method validation data
           obtained by the analyses of fortified tap water and raw source water
           are given  in  Tables  5-7.
     13.2 Demonstrated MDLs  are  given  in  Table  2.
          following equations were  used:
                                                   To  calculate  MDLs,  the
                 MDL = S t
                          (n-1,1-alpha = 0.99)
               where:
14.
                   t(n-i i-aipha = o 991 " Student's t value for the 99%
                   confidence level with n-1 degrees of  freedom

                   n = number of replicates

                   S = standard deviation of replicate analyses.

    POLLUTION PREVENTION

    14.1 One option of this method utilizes the new liquid-solid extraction
         (LSE) technology to remove the analytes from water.   It requires the


                                   506-19

-------
          use of very small  volumes of organic solvent and very small
          quantities of pure analytes, thereby eliminating the potential
          hazards to both the analyst and the environment.  The other  option
          in this method uses significant volumes of organic solvents.   It is
          highly recommended that laboratories use solvent recovery systems to
          recover used solvent as sample extracts are being concentrated.
          Recovered solvents should be recycled or properly disposed of.

     14.2 For information about pollution prevention that may be applicable to
          laboratory operations, consult "Less Is Better:  Laboratory Chemical
          Management for Waste Reduction" available from the American Chemical
          Society's Department of Government Relations and Science Policy,
          1155 16th Street N.W., Washington, D.C., 20036.

15.  WASTE MANAGEMENT

     15.1 It is the laboratory's responsibility to comply with all federal,
          state, and local regulations governing waste management,
          particularly the hazardous waste identification rules and land
          disposal restrictions.  The laboratory using this method has the
          responsibility to protect the air, water, and land by minimizing and
          controlling all releases from fume hoods and bench operations.
          Compliance is also required with any sewage.discharge permits and
          regulations.  For further information on waste management, consult
          "The Waste Management Manual for Laboratory Personnel," also
          available from the American Chemical Society at the address in Sect.
          14.2.

16.  REFERENCES

     1.   Glaser, J.V., D.L. Foerst,  G.D. McKee, S.A. Quave, and W.L. Budde,
          "Trace Analysis for Waste Waters," Environ. Sci. Techno!. 15, 1426,
          1981.

     2.   "Determination of Phthalates  in Industrial  and  Municipal
          Wastewaters,"  EPA-600/4-81-063, U.S.  Environmental Protection
          Agency, Environmental Monitoring and Support Laboratory, Cincinnati,
          Ohio 45268, October 1981.

     3.   Giam, C.S., Chan, H.S. and  Nef, G.S. "Sensitive Method  for
          Determination of  Phthalate  Ester  Plasticizers  in Open-Ocean Biota
          Samples," Anal. Chem.. 47,  2225 (1975).

     4.   Giam, C.S., and Chan, H.S.  "Control  of Blanks  in the  Analysis of
          Phthalates  in Air  and Ocean Biota  Samples," U.S. National Bureau of
          Standards,  Special  Publication 442,  pp.  701-708, 1976.

     5.   "Carcinogens - Working with Carcinogens,"  Department  of Health,
          Education,  and Welfare,  Public Health  Service,  Center for Disease
          Control,  National  Institute for Occupational Safety  and Health,
          Publication No. 77-206,  August 1977.


                                     506-20

-------
6.   "OSHA Safety and Health Standards, General Industry," (29 CFR 1910),
     Occupational Safety and Health Administration, OSHA 2206 (Revised,
     January 1976).                              ,

7.   "Safety in Academic Chemistry Laboratories," American Chemical
     Society  Publication, Committee on Chemical Safety, 3rd Edition
     1979.

8.   ASTM Annual Book of Standards, Part 31, D3694-78.  "Standard
     Practices for Preparation of Sample Containers and for Preservation
     of Organic  Constituents," American Society for Testing and
     Materials, Philadelphia.

9.   ASTM Annual Book of Standards^ Part 31, D3370. "Standard Practices
     for Sampling Water," American Society for Testing and Materials,
     Philadelphia.

10.   Munch,  J.  W., "Method 525.2-Determination of Organic Compounds in
     Drinking Water by Liquid-Solid Extraction and  Capillary Column
     Chromatography/ Mass Spectrometry" in Methods  for the Determination
     of Organic Compounds in Drinking Water; Supplement 3  (1995).
     USEPA,  National  Exposure Research Laboratory,  Cincinnati,  Ohio
     45268.
                              506-21

-------
17.   TABLES. DIAGRAMS. FLOWCHARTS.
 VALIDATION DATA
           TABLE  1.   RETENTION DATA AND CHROMATOGRAPHIC CONDITIONS
Parameter
Dimethyl phthalate
Di ethyl phthalate
Di-n-butyl phthalate
Butyl benzyl phthalate
Bis(2-ethylhexyl) adipate
Bis(2-ethylhexyl) phthalate
Di-n-octyl phthalate
Retention
(min)
Column 1
17.23
20.29
27.57
34.19
34.85
37.51
41.77
Time
Column 2
,17.89
21.13
28.67
35.34
36.76
39.58
44.44
        Column 1:  DB-5, fused silica capillary, 30 m x 0.32 mm I.D.,
        0.25 micron film thickness, Helium linear velocity = 30 cm/s.

        Column 2:  DB-1, fused silica capillary, 30 m x 0.32 mm I.D.,
        0.25 micron film thickness, Helium linear velocity = 30 cm/s.
        Chromatographic Conditions:
Injector temperature = 295°C
Detector temperature = 295°C
Program - 1 min hold at 60°C,
6°C/min to 260°C, 10 min hold.
Splitless injection  with 45 s
delay
                                      506-22

-------
TABLE 2.  ACCURACY, PRECISION, AND METHOD DETECTION LIMIT DATA FROM
          SIX LIQUID-LIQUID  EXTRACTION ANALYSES OF FORTIFIED REAGENT WATER
i
Analyte
Dimethyl phthalate
Diethyl phthalate
Di-n-butyl phthalate
Butyl benzyl phthalate
Bis(2-ethylhexyl) adipate
Bis(2-ethylhexyl) phthalate
Di-n-octyl phthalate
True
Cone.
M9/L
2.02
1.51
2.62
6.00
6.03
5.62
17.18
Mean
Meas.
Cone.
M9/L
1.42
1.16
1.78
3.27
• 3.94
2.92
7.96
.Mean
Std. Accuracy
Dev. % of True
M9/L Cone.
0.38
0.28
0.41
0.89
1.44
0.75
2.14
70.3
76.8
67.9
54.5
65.3
52.0
46.3
MDL
M9/L
1.14
0.84
1.23
2.67
11.82
2.25
6.42
                              506-23

-------
TABLE 3.  ACCURACY AND PRECISION DATA FROM SEVEN LIQUID-LIQUID
           EXTRACTION ANALYSES  OF  FORTIFIED  REAGENT  WATER
True
Concentration
Analyte M9/L
Dimethyl phthalate
Di ethyl phthalate
Di-n-butyl phthalate
Butyl benzyl phthalate
Bis(Z-ethylhexyl) adipate
Bis(2-ethylhexyl) phthalate
Di-n-octyl phthalate
15
15
15
15
30
30
30
Relative
Mean Accuracy Standard
% of True Deviation
Concentration %
73 16
71 16
68 15
71 15
69 18
67 21
.62 23
                              506-24

-------
TABLE 4.  ACCURACY AND PRECISION DATA FROM SIX LIQUID-SOLID
          EXTRACTION ANALYSES OF FORTIFIED REAGENT WATER
True
Concentration
Analyte Mg/L
Dimethyl phthalate
Diethyl phthalate
Di-n-butyl phthalate
Butyl benzyl phthalate
Bis(2-ethylhexyl) adipate
Bis(2-ethylhexyl) phthalate
Di-n-octyl phthalate
15
15
15
15
30
30
30
Relative
Mean Accuracy Standard
% of True Deviation
Concentration %
74
85
74
72
84
101
85
11 .
10
11
14
11
13
13
                           506-25

-------
        TABLE 5.   ACCURACY AND PRECISION DATA FROM SIX LIQUID-LIQUID
                   EXTRACTION ANALYSES OF FORTIFIED TAP WATER
Analyte
     True
Concentration
     M'9/L
                Relative
Mean Accuracy   Standard
 % of True      Deviation
Concentration      %
Dimethyl phthalate
Di ethyl phthalate
Di-n-butyl phthalate
Butyl benzyl phthalate
Bis(2-ethylhexyl) adipate
Bis(Z-ethylhexyl) phthalate
Di-n-octyl phthalate
5
5
5 '.
5
5
5
5
103
106
94
93
87
93
72
10.0 .
10.0
6.8
9.1
12.0
4.9
26.0
                                      506-26

-------
TABLE 6.  ACCURACY AND PRECISION DATA FROM SIX LIQUID-LIQUID
          EXTRACTION  ANALYSES  OF  FORTIFIED  RAW  SOURCE WATER
True
Concentration
Analyte ^g/L
Dimethyl phthalate
Di ethyl phthalate
Di-n-butyl phthalate
Butyl benzyl phthalate
Bis(2-ethylhexyl) adipate
Bis(2-ethylhexyl) phthalate
Di-n-octyl phthalate
' 5
5
5
5
5
5
5
Relative
Mean Accuracy Standard
% of True Deviation
Concentration %
59
78
99
72
115
91
54
51
45
29
23
32
35
24
                          506-27

-------
TABLE 7.  ACCURACY AND PRECISION DATA FROM SIX LIQUID-SOLID
          EXTRACTION  ANALYSES  OF FORTIFIED  RAW  SOURCE  WATER
True
Concentration
Analyte M9/L
Dimethyl phthalate
Di ethyl phthalate
Di-n-butyl phthalate
Butyl benzyl phthalate
Bis(2-ethylhexyl) adipate
Bis(Z-ethylhexyl) phthalate
Di-n-octyl phthalate
5
5
5
5
5
5
5
Mean Accuracy
% of True
Concentration
110
111
95
82
65
60
53
Standard
Deviation
%
20
32
30
20
24
21
15
                              506-28

-------
                      2 Uttr
                    separator^
                      funnel
                 125ml
                 solvent
                reservoir

               ground gfass T 14/3 5

               USE onndgt
               rubbef it op per

               No. 18-20IO«f-k>k
                   syringe needle
                     1 Bter
                 vacuum fl*sk
                                                         125ml
                                                         solvent
                                                        rtsfrvoir

                                                         ground gfass
                                                          ¥14/35   ••
                                                       LSH cartridge
                                                          100ml
                                                        separator/
                                                          funnel •
                   drying
                  column
                     ,
               1. 2 cm x 40 cm
                                                        10ml
                                                      graduated
                                                         vial
A. Extraction apparatus
                          ncvu  i
                             506-29
©. E!ut!on apparatus

-------
t3
oc
    11 h n 11111111111111 m i -n n 11111111 n n H it 11 n iu J i u > T i»11 * 11 T 11 * i

                           TIME (MIN.)
     Peaks obtained by Injecting S n9 for the  1st, 2nd, 4th and 5th
     coopounds, 10 ng for tht 6th, 7th thd 8th ewnpounds, ind 2.5 ng
     for the 3rd compound.  (Table 1)
                           F16URC 2
                            506-30

-------
      METHOD  507.     DETERMINATION OF NITROGEN- AND PHOSPHORUS-CONTAINING
                     PESTICIDES  IN WATER BY GAS CHROMATOGRAPHY WITH A NITROGEN-
                     PHOSPHORUS  DETECTOR
                                  Revision 2.1
                          Edited by J.W.  Munch (1995)
T. Engel (Battelle Columbus Laboratories) and D. Munch (U.S. EPA, Office of
Water), National Pesticide Survey Method 1, Revision 1.0 (1987)


R. L. Graves - Method 507, Revision 2.0 (1989)
                    NATIONAL EXPOSURE RESEARCH LABORATORY
                      OFFICE  OF  RESEARCH  AND  DEVELOPMENT
                    U.S.  ENVIRONMENTAL PROTECTION  AGENCY
                           CINCINNATI, OHIO  45268
                                    507-1

-------
                                  METHOD 507

   DETERMINATION OF NITR06EN-AND PHOSPHORUS-CONTAINING PESTICIDES IN WATER
           BY GAS  CHROMATOGRAPHY WITH A NITROGEN-PHOSPHORUS DETECTOR
1.   SCOPE AND APPLICATION

     1.1  This is a gas chromatographic (GC) method applicable to the
          determination of certain nitrogen- and phosphorus-containing
          pesticides in ground water and finished drinking water.  The
          following compounds can be determined using this method:
               Analvte

               Alachlor
               Ametryn
               Atraton
               Atrazine
               Bromacil
               Butachlor
               Butyl ate
               Carboxin
               Chlorpropham
               Cycloate
               Diazinon(a)*
               Dichlorvos
               Diphenamid
               Disulfoton*
               Disulfoton  sulfone*
               Disulfoton  sulfoxide(a)'
               EPIC
               Ethoprop
               Fenamiphos
               Fenarimol
               Fluridone
               Hexazinone
               Merphos*
               Methyl paraoxon
               Metolachlor
               Metribuzin
               Mevinphos
               MGK 264
               Molinate
               Napropamide
               Norflurazon
               Pebulate
               Prometon
               Prometryn
               Pronamide(a)*
               Propazine
               Simazine
               Simetryn
               Stirofos
               Tebuthiuron
Chemical Abstract Services
 	Registry Number	
        15972-
          834-
         1610-
         1912-
          314-
        23184-
         2008-
         5234-
          101-
         1134-
          333-
           62-
          957-
          298-
         2497-
         2497-
          759-
        13194-
        22224-
        60168-
        59756-
        51235-
          150-
          950-
        51218-
        21087-
         7786-
          113-
         2212-
        15299-
        27314-
         1114-
         1610-
         7287-
        23950-
          139-
          122-
         1014-
        22248-
        34014-
60-8
12-8
17-9
24-9
40-9
66-9
41-5
68-5
21-3
23-2
41-5
73-7
51-7
04-4
06-5
07-6
94-4
48-4
92-6
88-9
60-4
04-2
50-5
35-6
45-2
64-9
34-7
48-4
67-1
99-7
13-2
71-2
18-0
19-6
58-5
40-2
34-9
70-6
79-9
18-1
                                     507-2

-------
                Terbacil                             5902-51-2
                Terbufos(a)*                        13071-79-9
                Terbutryn                             886-50-0
                Triademefon                         43121-43-3
                Tricyclazole        .                41814-78-2
                Vernolate                            1929-77-7

           (a)  Compound exhibits aqueous instability.   Samples for which this
                compound is an analyte of interest must be extracted
                immediately (Sections  11.1 through 11.3).

             *   These compounds are only qualitatively  identified.  These
                compounds are not quantitated because  control  over precision
                has not been accomplished.

      1.2  This method  has  been  validated  in  a  single laboratory and estimated
          detection  limits  (EDLs)  and method detection limits  (MDLs) have been
          determined for the  analytes above  (Sect.  13).   Observed  detection
          limits may vary  among  waters, depending  upon the nature  of
          interferences  in  the  sample matrix and  the specific  instrumentation
          used.

      1.3  This method  is restricted to  use by  or  under the supervision of
          analysts experienced  in  the use of GC and in the interpretation of
         'gas chromatograms.   Each analyst must demonstrate the ability to
          generate acceptable  results with this method using the procedure
          described  in  Sect.  9.

      1.4  Analytes that are not  separated chromatographically, i.e., analytes
          which have very  similar  retention times, cannot be individually
          identified and measured  in the  same  calibration mixture  or water
          sample unless an  alternative  technique  for identification and.
          quantitation  exist  (Section 11.5).

      1.5  When this method  is used to analyze  unfamiliar  samples for any or
          all of the analytes above, analyte identifications should be
          confirmed by  at  least  one additional qualitative technique.

2.   SUMMARY OF METHOD

     2.1  A measured volume of sample of approximately 1  L is extracted with
          methylene ch'loride by  shaking in a separatory funnel or mechanical
          tumbling in a bottle.  The methylene chloride extract is isolated,
          dried and concentrated to a volume of 5 ml during a solvent exchange
          to methyl tert-butyl ether (MTBE).  Chromatographic conditions are
          described which permit the separation and measurement of the
          analytes in the extract by Capillary Column GC with a nitrogen-
          phosphorus detector (NPD).

3.   DEFINITIONS

     3.1  INTERNAL STANDARD — A pure analyte(s)  added  to a solution in known
          amount(s) and used to measure the relative responses of other method
          analytes and surrogates that are components of the same solution.
          The internal  standard must be an analyte that is not a sample
          component.

                                    507-3

-------
3.2  SURROGATE ANALYTE — A pure analyte(s), which is extremely unlikely
     to be found in any sample, and which is added to a sample aliquot in
     known amount(s) before extraction and is measured with the same
     procedures used to measure other sample components.  The purpose of
     a surrogate analyte is to monitor method performance with each
     sample.

3.3  LABORATORY DUPLICATES (LD1 and LD2) — Two sample aliquots taken in
     the analytical laboratory and analyzed separately with Identical
     procedures.  Analyses of LD1 and LD2 give a measure of the precision
     associated with laboratory procedures, but not with sample
     collection, preservation, or storage procedures.

3.4  FIELD DUPLICATES (FD1 and FD2) —Two separate samples collected at
     the same time and place under identical circumstances and treated
     exactly the same throughout field and laboratory procedures.
     Analyses of FD1 and FD2 give a measure of the precision associated
     with sample collection, preservation and storage,"as well as with
     laboratory procedures.

3.5  LABORATORY REAGENT BLANK  (LRB) — An aliquot of reagent water that
     is treated exactly as a sample including exposure to all' glassware,-
     equipment, solvents, reagents, internal standards, and surrogates
     that are used with other samples.  The LRB is used to determine if
     method analytes or other interferences are present in the laboratory
     environment, the reagents, or the apparatus.

3.6  FIELD REAGENT BLANK (FRB) — Reagent water placed in a sample
     container in the laboratory and treated as a sample in all respects,
     including exposure to sampling site conditions, storage, preserva-
     tion and all analytical procedures.  The purpose of the FRB is to
     determine if method analytes or other1interferences are present in
     the field environment.

3.7  LABORATORY PERFORMANCE CHECK SOLUTION  (LPC) — A solution of method
     analytes, surrogate compounds, and internal standards used to
     evaluate the performance of the instrument system with respect to a
     defined set of method criteria.

3.8  LABORATORY FORTIFIED BLANK (LFB) — An aliquot of reagent water to
     which known quantities of the method analytes are added in the
     laboratory.  The LFB is analyzed exactly like a sample, and its
     purpose is to determine whether the methodology is in control, and
     whether the laboratory is capable of making accurate and precise
     measurements at the required method detection limit.

3.9  LABORATORY FORTIFIED SAMPLE MATRIX (LFM) — An aliquot of an
     environmental sample to which known quantities of the method
     analytes are added in the laboratory.  The LFM is analyzed exactly
     like a sample, and its purpose is to determine whether the sample
     matrix contributes bias to the analytical results. The background
     concentrations of the analytes in the  sample matrix must be
     determined in a separate  aliquot and the measured values in the LFM
     corrected for background concentrations.
                                507-4

-------
     3.10 STOCK STANDARD SOLUTION — A concentrated solution containing a
          single certified standard that is a method analyte, or a
          concentrated solution of a single analyte prepared in the laboratory
          with an assayed reference compound.  Stock standard solutions are
          used to prepare primary dilution standards.

     3.11 PRIMARY DILUTION STANDARD SOLUTION — A solution of several analytes
          prepared in the laboratory from stock standard solutions and diluted
          as needed to prepare calibration solutions and other needed analyte
          solutions.

     3.12 CALIBRATION STANDARD (CAL) — A solution prepared from the primary
          dilution standard solution and stock standard solutions of the
          internal standards and surrogate analytes.  The CAL solutions are
          used to calibrate the instrument response with respect to analyte
          concentration.

     3.13 QUALITY CONTROL SAMPLE (QCS) — A sample matrix containing method
          analytes or a solution of method analytes in a water miscible
          solvent which is used to fortify reagent water or environmental
          samples.  The QCS is obtained from a source external  to the
          laboratory, and is used to check laboratory performance with
          externally prepared test materials.

4.   INTERFERENCES

     4.1  Method interferences may be caused by contaminants in solvents,
         . reagents,  glassware and other sample processing apparatus that lead
          to discrete artifacts or elevated baselines in gas chromatograms.
          All reagents and apparatus must be routinely demonstrated to be  free
          from interferences under the conditions of the analysis by running
          laboratory reagent blanks as described in Sect.  9.2.

          4.1.1  Glassware must be scrupulously cleaned (1).   Clean all  glass-
                 ware as soon as possible after use by thoroughly rinsing  with
                 the last solvent used in it.  Follow by washing with hot
                 water and detergent and thorough rinsing with  tap and reagent
                 water.   Drain dry,  and heat in an oven or muffle furnace  at
                 400°C for 1  hour.   Do not heat volumetric ware.   Thermally
                 stable materials might not be eliminated by this treatment.
                 Thorough rinsing with acetone may be substituted for the
                 heating.   After drying and cooling,  seal  and store glassware
                 in  a clean environment to prevent any accumulation of dust or
                 other contaminants.   Store inverted or capped  with aluminum
                 foil.

          4.1.2  The use of high  purity reagents  and solvents helps to
                 minimize interference problems.   Purification  of solvents by
                 distillation in  all-glass  systems may be  required.   WARNING:
                 When a  solvent is  purified,  stabilizers  added  by the
                 manufacturer may be  removed thus  potentially making  the
                 solvent hazardous.   Also,  when a  solvent  is  purified,
                 preservatives added  by the manufacturer  are  removed  thus
                 potentially  reducing  the shelf-life.
                                    507-5

-------
     4.2  Interfering contamination may occur when a sample containing low
          concentrations of analytes is analyzed immediately following a
          sample containing relatively high concentrations of analytes.
          Between-sample rinsing of the sample syringe and associated
          equipment with MTBE can minimize sample cross contamination.  After
          analysis of a sample containing high concentrations of analytes, one
          or more injections of MTBE should be made to ensure that accurate
          values are obtained for the next sample.

     4.3  Matrix interferences may be caused by contaminants that are
          coextracted from the sample.  Also, note that all the analytes
          listed in the scope and application section are not resolved from
          each other on any one column, i.e., one analyte of interest may be
          an interferant for another analyte of interest.  The extent of
          matrix interferences will vary considerably from source to source,
          depending upon the water sampled.  Further processing of sample
          extracts may be necessary.  Positive identifications should be
          confirmed (Sect. 11.5).

     4.4  It is important that samples and working standards be contained in
          the same solvent.  The solvent for working standards must be the
          same as the final solvent used in sample preparation.  If this is
          not the case, chromatographic comparability of standards to sample
          may be affected.

5.   SAFETY

     5.1  The toxicity or carcinogenicity of each reagent used in this method
          has not been precisely defined; however, each chemical compound must
          be treated as a potential health hazard.  Accordingly, exposure to
          these chemicals must be reduced to the lowest possible level.   The
          laboratory is responsible for maintaining a current awareness file
          of OSHA regulations regarding the safe handling of the chemicals  ~
          specified in this method.  A reference file of material safety data
          sheets should also be made available, to all personnel involved in
          the chemical analysis.  Additional references to laboratory safety
          are available and have been identified (2-4) for the information of
          the analyst.

     5.2  WARNING:  When a solvent is purified, stabilizers added by the
          manufacturer may be removed thus potentially making the solvent
          hazardous.

6.   EQUIPMENT AND SUPPLIES  (All specifications are suggested.  Catalog
     numbers are included for illustration only.)

     6.1  Sample bottle — Borosilicate, 1-L volume with graduations  (Wheaton
          Media/Lab bottle 219820 or equivalent), fitted with screw caps lined
          with TFE-fluorocarbon.  Protect samples from light.  Amber bottles
          may be used.  The container must be washed and dried as described in
          Sect. 4.1.1 before use to minimize contamination.  Cap liners are
          cut to fit from sheets (Pierce Catalog No. 012736 or equivalent) and
          extracted with methanol overnight prior to use.
                                     507-6

-------
6.2  GLASSWARE

     6.2.1  Separatory  funnel -- 2000-ml, with TFE-fluorocarbon stopcock,
            ground glass or TFE-fluorocarbon stopper.

     6.2.2  Tumbler bottle -- 1.7-L  (Wheaton Roller Culture. Vessel or
            equivalent), witn TFE-fluorocarbon lined screw cap.  Cap
            liners are  cut to fit from sheets  (Pierce Catalog No. 012736)
            and extracted with methanol overnight prior to use.

     6.2.3  Flask, Erlenmeyer -- 500-mL.

     6.2.4  Concentrator tube, Kuderna-Danish  (K-D) — 10- or 25-mL,
            graduated (Kontes K-570050-2525 or K-570050-1025 or
            equivalent).  Calibration must be checked at, the volumes
            employed in the test.  Ground glass stoppers are used to
            prevent evaporation of extracts.

     6.2.5  Evaporative flask, K-D — 500-mL (Kontes K-570001-0500 or
            equivalent).  Attach to  concentrator tube with springs.

     6.2.6  Snyder column, K-D -- Three-ball macro (Kontes K-503000-0121
            or equivalent).

     6.2.7  Snyder column, K-D -- Two-ball micro (Kontes K-569001-0219 or
            equivalent).

     6.2.8  Vials — glass, 5- to 10-mL capacity with TFE-fluorocarbon
            1ined screw cap.                                 ,

6.3  Separatory funnel  shaker (Optional) -- Capable of holding 2-L
     separatory funnels and shaking  them with rocking motion to achieve
     thorough mixing of separatory funnel contents (available from
     Eberbach Co. in Ann Arbor,  MI or other suppliers).

6.4  Tumbler — Capable of holding tumbler bottles and tumbling 'them
     end-over-end at 30 turns/min (Associated Design and Mfg. Co.,
     Alexandria, VA. or other suppliers).

6.5  Boiling stones — Carborundum,   #12 granules (Arthur H. Thomas Co.
     #1590-033 or equivalent).  Heat at 400°C for 30 min prior to use.
     Cool and store in desiccator.

6.6  Water bath — Heated, capable of temperature control  (+ 2°C).  The
     bath should be used in a hood.

6.7  Balance— Analytical,  capable  of accurately weighing to the nearest
     0.0001 g.

6.8  GAS CHROMATOGRAPH — Analytical  system complete with  temperature
     programmable GC suitable for use with capillary columns and all
     required accessories including  syringes,  analytical  columns,  gases,
     detector and stripchart recorder.  A data system is recommended for
     measuring peak areas.  Table 1   lists retention times  observed for
     method analytes using the columns and analytical  conditions
     described below.

      ;    '                     507-7

-------
6.8.1
6.8.2
                 Column  1  (Primary column) — 30 m long x 0.25 mm I.D. DB-5
                 bonded_fused  silica column, 0.25 fim film thickness (J&W
                 Scientific) or equivalent.  Helium carrier gas flow is
                 established at 30 cm/sec linear velocity and oven temperature
                 is programmed from 60°C to 300°C at 4°C/min.  Data presented
                 in this method were obtained using this column.  The
                 injection volume was 2 ni in splitless mode with a 45 s
                 delay.  The injector temperature was 250°C and the detector
                 temperature was 300°C.  Alternative columns may be used in
                 accordance with the provisions described in Sect. 9.4.

                 Column  2  (Confirmation column) — 30 m long x 0.25 mm
                 I.D.DB-1701 bonded fused silica column, 0.25 urn film
                 thickness (J&W Scientific) or equivalent.  He!ium. carrier gas
                 flow is established at 30 cm/sec linear velocity and oven
                 temperature is programmed from 60C to 300°C at 4°C/min.

          6.8.3  Detector — Nitrogen-phosphorus (NPD).

7.   REAGENTS AND STANDARDS — WARNING:  When a solvent  is purified,
     stabilizers added by the manufacturer are removed thus potentially making
     the solvent hazardous.  Also,  when a solvent is purified,  preservatives
     added by the manufacturer are removed thus potentially reducing  the
     shelflife.

     7.1  Acetone,  methylene chloride,  methyl  tert.-butyl  ether (MTBE)  —
          Distilled-in-glass quality or equivalent.

     7.2  Phosphate buffer, pH 7 — Prepare by mixing 29.6 ml 0.1 N HC1 and 50
          ml 0.1 M dipotassium phosphate.

     7.3  Sodium chloride (NaCl),  crystal,  ACS grade —  Heat treat in  a
          shallow tray at 400°C for a minimum of 4 hours to remove interfering
          organic substances.   Store in a  glass bottle (not plastic)  to avoid
          phthalate contamination.

     7.4  Sodium sulfate, granular,  anhydrous,  ACS grade — Heat treat  in a
          shallow tray at 400°C for a minimum of 4 hours to remove interfering
          organic substances.   Store in a  glass bottle (not plastic)  to avoid
          phthalate contamination.

     7.5  Sodium thiosulfate,  granular,  anhydrous,  ACS grade.

     7.6  Triphenylphosphate  (TPP)  -- 98%  purity,  for use  as internal  standard
          (available from Aldrich  Chemical  Co.).

     7.7  l,3-Dimethyl-2-nitrobenzene — 98%; purity,  for use as  surrogate
          standard  (available  from  Aldrich  Chemical  Co.).

     7.8  Mercuric  Chloride —  ACS  grade (Aldrich  Chemical  Co.), -  for  use as a
          bactericide  (optional-  see  Sect.  8).

     7.9  Reagent water  — Reagent  water is defined  as a water  that is
          reasonably free of contamination  that  would  prevent the
          determination  of any  analyte  of  interest.   Reagent water used to
                          507-8

-------
                                                                                      I
     generate the validation data in this method was distilled water
     obtained from the Magnetic Springs Water Co., Columbus, Ohio.

7.10 STOCK STANDARD SOLUTIONS (1.00 fig/pi) — Stock standard solutions
     may be purchased as certified solutions or prepared from pure
     standard materials using the following procedure:

     7.10.1 Prepare stock standard solutions by accurately weighing
            approximately 0.0100 g of pure material.  Dissolve the
            material in MTBE and dilute to volume in a 10-mL volumetric
            flask.  The stock solution for simazine should be prepared in
            methanol.  Larger volumes may be used at the convenience of
            the analyst.  If compound purity is certified at 96% or
            greater, the weight may be used without correction to
            calculate the concentration of the stock standard.
            Commercially prepared stock standards may be used at any
            concentration if they are certified by the manufacturer or by
            an independent source.

     7.10.2 Transfer the stock standard solutions into
            TFE-fluorocarbon-sealed screw cap amber vials.  Store at room
            temperature and protect from light.                          .

     7.10.3 Stock standard solutions should be replaced after two months
            or sooner if comparison with laboratory fortified blanks, or
            QC samples indicate a problem.

7.11 INTERNAL STANDARD SOLUTION — Prepare the internal standard solution
    , by accurately weighing approximately 0.0500 g of pure TPP.  Dissolve
     the TPP in MTBE and dilute to volume in a 100-mL volumetric flask.
     Transfer the internal standard solution to a TFE-fluorocarbon-sealed
     screw cap bottle and store at room temperature.  Addition of 50 /iL
     of the internal standard solution to 5 mL of sample extract results
     in a final TPP concentration of 5.0 /xg/mL.  This solution should be
     replaced when ongoing QC (Sect. 9) indicates a problem.  Note that
     TPP has been shown to be an effective internal  standard for the
     method analytes, but other compounds may be used if the quality
     control, requirements in Sect. 9 are met.

7.12 SURROGATE STANDARD SOLUTION — Prepare the surrogate standard
     solution by accurately weighing approximately 0.0250 g of pure
     l,3-dimethyl-2-nitrobenzene.  .Dissolve the 1,3-dimethyl-
     2-nitrobenzene in MTBE and dilute to volume in a 100-mL volumetric
     flask.  Transfer the surrogate standard solution to a
     TFE-fluorocarbon-sealed screw cap bottle and store at room
     temperature.   Addition of 50 /uL of the surrogate standard solution
     to a 1-L sample prior to extraction results in a 1,3-dimethyl-
     2-nitrobenzene concentration in the sample of 12.5 /ig/L.  Solution
     should be replaced when ongoing QC (Sect.  9) indicates a problem.
     Note that l,3-dimethyl-2-nitrobenzene has been shown to be an
     effective surrogate standard for the method analytes, but other
     compounds may be used if the qua!ity control requirements in Sect. 9
     are met-.

7.13 LABORATORY PERFORMANCE CHECK SOLUTION —' Prepare the laboratory
     performance check solution by adding 5 [j,L of the vernolate stock

                                507-9

-------
          solution, 0.5 mL of the bromacil stock solution, 30 /ul_ of the
          prometon stock solution, 15 /iL of the atrazine stock solution,  1.0
          ml of the surrogate solution, and 500 nl of the internal standard
          solution to a 100-mL volumetric flask.  Dilute to volume with MTBE
          and thoroughly mix the solution.  Transfer to a TFE-fluorocarbon-
          sealed screw cap bottle and store at room temperature.  Solution
          should be replaced when ongoing QC (Sect. 9) indicates a problem.

8.   SAMPLE COLLECTION, PRESERVATION. AND STORAGE

     8.1  Grab samples must be collected in glass containers.  Conventional
          sampling practices (5) should be followed; however, the bottle must
          not be prerinsed with sample before collection.

     8.2  SAMPLE PRESERVATION AND STORAGE

          8.2.1  If residual chlorine is present, add 80 mg of sodium
                 thiosulfate per liter of sample to the sample bottle prior to
                 collecting the sample.

          8.2.2  After the sample is collected in a bottle containing sodium
                 thiosulfate, seal the bottle and shake until dissolved.

          8.2.3  The samples must be iced or refrigerated at 4°C away from
                 light from the time of collection until extraction.  Pre-
                 servation study results indicated that most method analytes
                 present in samples were stable for 14 days when stored under
                 these conditions.  The analytes disulfoton sulfoxide,
                 diazinon, pronamide, and terbufos exhibited significant
                 aqueous instability, and samples to be analyzed for these
                 compounds must be extracted immediately.  The analytes
                 carboxin, EPTC, fluridone, metolachlor, napropamide,
                 tebuthiuron, and terbacil exhibited recoveries of less than
                 60% after 14 days.  Analyte stability may be affected by the
                 matrix; therefore, the analyst should verify that the
                 preservation technique is applicable to the samples under
                 study.

          8.2.4  All performance data presented in this method are from
                 samples preserved with mercuric chloride.  No suitable
                 preservation agent (biocide) has been found other than
                 mercuric chloride.  However the use of mercuric chloride is
                 not required due to its toxicity and potential harm to the
                 environment.

          8.2.5  In some circumstances where biological degradation of target
                 pesticides might be expected, use of mercuric chloride may be
                 appropriate to minimize the possibility of false-negative
                 results.  If mercuric chloride is to be used, add it (See
                 7.8) to the sample bottle in amounts to produce a
                 concentration of 10 mg/L.  Add 1 ml of a solution containing
                 10 mg/ml of mercuric chloride in reagent water to the sample
                 bottle at the sampling site or in the laboratory before
                 shipping to the sampling site.  A major disadvantage of
                 mercuric chloride is that it is a highly toxic chemical;


                                    507-10

-------
                 mercuric chloride must be handled with caution,  and samples
                 containing mercuric chloride must be disposed of properly.

     8.3  Extract Storage -- Extracts should be stored at 4°C away from light.
          Preservation study results indicate that most analytes  are stable
          for 28 days; however, a 14-day maximum extract storage  time is
          recommended.  The analyst should verify appropriate extract holding
          times applicable to the samples under study.

9.   QUALITY CONTROL

     9.1  Minimum quality control (QC) requirements are  initial  demonstration
          of laboratory capability, determination of surrogate compound
          recoveries in each sample and blank, monitoring internal standard
          peak area or height in each sample and blank (when internal standard
          calibration procedures are being employed), analysis of laboratory
          reagent blanks, laboratory fortified samples, laboratory fortified
          blanks, and QC samples.  A method detection limit (MDL) must also be
          determined for each analyte.

     9.2  Laboratory Reagent Blanks.  Before processing any samples, the
          analyst must demonstrate that all glassware and reagent
          interferences are under control.  Each time a set of samples is
          extracted or reagents are changed, a LRB must be analyzed.  If
          within the retention time window of any analyte of interest the LRB
          produces a peak that would prevent the determination of that
          analyte, determine the source of contamination and eliminate the
          interference before processing samples.

     9.3  Initial Demonstration of Capability.

          9.3.1  Select a representative fortified concentration  (about 10
                 times EDL or at a concentration in the middle of the
                 calibration range established in Sect. 10) for each analyte.
                 Prepare a standard concentrate containing each analyte at
                 1000 times the selected concentration.  With a syringe, add 1
                 mL of the concentrate to each of four to seven 1-L aliquots
                 of reagent water, and analyze each aliquot according to
                 procedures beginning in Sect. 11.

          9.3.2  For each analyte, the mean recovery value for these samples
                 must fall in the range of R ±30% using the values for R for
                 reagent water in Table 2.  The RSD for these measurements
                 must be 20% or less.  For those compounds that meet the
                 acceptance criteria, performance is considered acceptable.
                 For those compounds that fail these criteria, this procedure
                 must be repeated using fresh replicate samples until
                 satisfactory performance has been demonstrated.

          9.3.3  For each analyte, determine the MDL.  Prepare a  minimum of 7
                 LFBs at a low concentration.  The fortification  concentration
                 in Table 3 may be used as a guide, or use calibration data
                 obtained in Section  10 to estimate a  concentration for each
                 analyte that will produce a peak with a 3-5 times  signal to
                 noise response.  Extract and analyze  each replicate according
                 to Sections 11 and  12.   It  is recommended that these LFBs  be

                                     507-11

-------
             prepared  and  analyzed  over a  period  of several  days,  so  that
             day to  day  variations  are  reflected  in the  precision
             measurement.   Calculate  mean  recovery  and standard  deviation
             for each  analyte.  Use  the  equation given  in  Table 3 to
             calculate the MDL.

      9.3.3   The initial demonstration  of  capability is  used  primarily to
             preclude  a  laboratory  from analyzing unknown  samples  via a
             new,  unfamiliar  method prior  to  obtaining some  experience
             with  it.  It  is  expected that  as  laboratory  personnel gain
             experience  with  this method the  quality of  data  will  improve
             beyond  those  required  here.

9.4   The  analyst  is permitted  to modify GC columns, GC  conditions,
      concentration  techniques  (i.e.  evaporation  techniques), internal
      standards  or surrogate  compounds.  Each  time  such method
      modifications  are  made, the analyst must repeat the  procedures  in
      Sect. 9.3.

9.5   Assessing  Surrogate  Recovery

      9.5.1   When  surrogate recovery  from a sample  or method  blank is <70%
             or  >130%, check  calculations to  locate possible  errors,
             fortifying  solutions for degradation,  contamination,  and
             instrument  performance.  If those steps do not reveal the
             cause of  the  problem, reanalyze  the extract.

      9.5.2   If  a  LRB  extract reanalysis fails the  70-130% recovery
             criterion,  the problem must be identified and corrected
             before continuing.

      9.5.3   If  sample extract reanalysis meets the surrogate recovery
             criterion,  report only data for the reanalyzed extract.  If
             sample extract reanalysis  continues to fail  the  recovery
             criterion,  report all data for that sample as suspect.

9.6  Assessing  the Internal Standard

     9.6.1  When using  the internal  standard calibration procedure, the
             analyst is  expected to monitor the IS response (peak area or
            peak height) of all samples during each analysis day.  The IS
            response for any sample  chromatogram should not deviate from
            the daily calibration check standard's IS response by more
            than 30%.

     9.6.2   If >30% deviation occurs with an individual  extract, optimize
            instrument performance and inject a second aliquot of that
            extract.

            9.6.2.1   If the reinjected aliquot produces  an acceptable
                     internal standard response report results for that
                     aliquot.

            9.6.2.2   If a  deviation of greater than 30% is obtained for
                     the reinjected extract,  analysis  of the sample
                     should be repeated beginning with Sect. 11, provided

                               507-12

-------
                     the sample is still available. Otherwise, report
                     results obtained from the reinjected extract, but
                  • .  annotate as suspect.

     9.6.3  If consecutive samples fail the IS response acceptance
            criterion, immediately analyze a calibration check standard.

            9.6.3.1  If the check standard provides a response factor
                     (RF) within 20% of the predicted value, then follow
                     procedures itemized in Sect. 9.6.2 for each sample
                     failing the IS response criterion.

            9.6.3.2  If the check standard provides a response factor
                     which deviates more than 20% of the predicted value,
                     then the analyst must recalibrate, as specified-in
                     Sect. 9.

9.7  Assessing Laboratory Performance - Laboratory Fortified Blank

     9.7.1  The laboratory must analyze at least one laboratory fortified
            blank (LFB) sample with every twenty samples or one per
            sample set (all samples extracted within a 24-h period)
            whichever is greater.   Ideally, the fortified concentration
            of each analyte in the LFB should be the same concentration
            selected in Section 9.3.1. '.Calculate accuracy as percent
            recovery (X,-)..   If the  recovery of,any analyte falls  outside
            the control limits (see Sect. 9.7.2), that analyte is judged
            out of control, and the source of the problem should be
            identified and resolved before continuing analyses.

     9.7.2  Until  sufficient data become available from within their own
            laboratory, usually a minimum of results from 20 to 30
            analyses, the laboratory should assess laboratory performance
          '  against the control  limits in Sect.  9.3.2 that are derived
            from the data in Table 2.  When sufficient internal
            performance data becomes available,  develop control limits
            from the mean percent recovery (X)  and standard deviation (S)
            of the percent recovery.  These data are used to establish
            upper and lower control  limits as follows:

                     UPPER CONTROL LIMIT  = X + 3S

                     LOWER CONTROL LIMIT =  X.- 3S

            After each five to ten new recovery measurements, new control
            limits should be calculated using only the most recent 20-30
            data points.   These calculated control limits must not exceed
            the fixed limits listed in Sect.  9.3.2.

     9.7.5  It is recommended that the laboratory periodically determine
            and document its detection limit capabilities for analytes of
            interest.

     9.7.6  At least  quarterly,  analyze a QC sample from an outside
            source.
                               507-13

-------
      9.8  Assessing Analyte Recovery - Laboratory Fortified Sample Matrix

           9.8.1  The laboratory must add a known concentration to a minimum of
                  5% of the routine samples or one sample per set, whichever is
                  greater.  The fortified concentration should not be less than
                  the background concentration of the sample selected for
                  fortification.  Ideally, the concentration should be the same
                  as that used for the laboratory fortified blank (Sect. 9.7).
                  Over time, samples from all  routine sample sources should be
                  fortified.

           9.8.2  Calculate the percent recovery, P,  of the concentration for
                  each analyte, after correcting the analytical  result,  X,  from
                  the fortified sample for the background concentration, b,
                  measured in the unfortified  sample, i.e.,:

                           P = 100 (X - b) / .fortifying concentration,

                  and compare these values to  reagent water recoveries  listed
                  in Table 2.   The calculated  value  of P must fall  in  the range
                  of R ± 35%.   If P exceeds  this control  limit  the results  for
                  that analyte in the unfortified matrix must be  listed  as
                  suspect due to matrix interference.

      9.9   ASSESSING INSTRUMENT SYSTEM -  LABORATORY  PERFORMANCE CHECK  (LPC)  -
           After initial  demonstration of  capability, instrument  performance
           should be monitored  on  a  daily  basis  by analysis  of  the  LPC sample.
           The  LPC sample contains  compounds designed to  monitor  instrument
           sensitivity,  column  performance (primary column)  and chromatographic
           performance.   LPC  sample .components and performance  criteria  are
           listed in Table 4.   Inability to  demonstrate  acceptable  instrument
           performance  indicates  the  need  for  reevaluation of the  instrument
           system.   The  sensitivity  requirements  are  set  based on  the EDLs
           published in  this method.   The  purpose of  the  sensitivity
           requirement  is  to monitor  the stability of instrument sensitivity,
           not  as  an absolute sensitivity  requirement.   If laboratory EDLs
           differ  from those listed  in this method, concentrations of the LPC
           standard  compounds must be  adjusted to be  compatible with the
           laboratory EDLs.

    9.10   The  laboratory  may adopt additional  quality control practices for '
           use  with  this method.  The specific practices that are most
           productive depend upon the needs of the laboratory and the nature of
           the,  samples.  For example, field or laboratory duplicates may be
           analyzed  to assess the precision of the environmental measurements
           or field  reagent blanks may be  used to  assess contamination of
           samples under site conditions,   transportation and storage.

10.   CALIBRATION

     10.1  Establish GC operating parameters equivalent to those indicated in
          Sect. 6.8.  The GC system may be calibrated using either the
          internal standard technique (Sect. 10.2) or the external standard
          technique (Sect. 10.3).  Be aware that NPDs may exhibit instability
          (i.e., fail to hold calibration curves over time).  The analyst may,
          when  analyzing samples for target analytes  which are very rarely

                                    507-14

-------
     found, prefer to analyze on a daily basis a low level (e.g. 5 to 10
     times detection limit or 1/2 times the regulatory limit,  whichever
     is less), sample (containing all analytes of interest) and require
     some minimum sensitivity (e.g.  1/2 full scale deflection) to show
     that if the analyte were present it would be detected.  The analyst
     may then quantitate using single point calibration (Sect. 10.2.5 or
     10.3.4).  NOTE:  Calibration standard solutions must be prepared
     such that no unresolved analytes are mixed together.

10.2 INTERNAL STANDARD CALIBRATION PROCEDURE — To usejthis approach, the
     analyst must select one or more internal  standards compatible in
     analytical  behavior to the compounds of interest.   The analyst must
     further demonstrate that the measurement of the internal  standard is
     not affected by method or matrix interferences.  TPP has  been
     identified  as a suitable internal  standard.

     10.2.1 Prepare calibration standards at a minimum of three
            (recommend five) concentration levels for each analyte of
            interest by adding volumes  of one or more stock standards to
            a volumetric flask.   Guidance on the number of standards is
            as follows:   A minimum of three calibration standards are
            required to calibrate a  range of a factor of 20 in
            concentration.   For a factor of 50 use at least four
            standards,  and for a factor of 100 at least five standards.
            The  lowest standard should  represent analyte concentrations
            near,  but above, their respective EDLs.   The remaining
            standards should bracket the analyte concentrations expected
            in the sample extracts,  or  should define the working range of
            the  detector.  If Merphos is to be determined, calibrate with
            DEF  (S,S,S-tributylphosphoro-trithioate) .   Merphos is
            converted to S,S,S-tributylphosphoro-trithioate (DEF) in the
            hot  GC injection port; DEF  is actually detected using the
            analysis conditions  in this method.   To  each calibration
            standard,  add a known constant amount of one or more of the
            internal  standards,  and  dilute to  volume with MTBE.

     10.2.2 Analyze each calibration standard  according to the procedure
            described in Sect.  11.4.  Tabulate response (peak  height or
            area)  against concentration for each compound and  internal
            standard.   Calculate the response  factor (RF)  for  each
            analyte and  surrogate using Equation 1.   RF is a unitless
            value.
            RF  =      sis                Equation  1

                   (Af.)(C.)

            where ":

            As  = Response for the analyte.

            Ajs = Response for the internal standard.

            Cjs = Concentration of the internal  standard

            Cs  = Concentration of the analyte to be measured M9/L.

                              507-15

-------
          10.2.3  If the  RF  value  over  the working  range  is  constant  (20%  RSD   > s;
                  or less) the  average  RF can  be  used  for calculations.       •    '
                  Alternatively, the  results can  be used  to  plot  a  calibration  , •
                  curve of response ratios  (As/Ajs)  vs.  Cs.                     '•  ;';

          10.2.4  The working calibration curve or  calibration  factor must be
                  verified on each working day by the  measurement of  a minimum
                  of two  calibration  check standards,  one at the  beginning and
                  one at  the end of the analysis  day.   These check  standards
                  should  be  at  two different concentration levels to  verify the
                  calibration curve.  For extended  periods of analysis (greater
                  than 8  hrs.), it is strongly recommended that check standards
                  be interspersed  with  samples at regular intervals during the
                  course  of  the analyses.   If  the response for  any  analyte
                  varies  from the  predicted response by more than ±20%,  the
                  test must  be  repeated using  a fresh  calibration standard.   If
                  the results still do  not agree, generate a new  calibration
                  curve.  For those analytes that failed  the calibration
                  verification, results from field  samples analyzed since  the
                  last passing  calibration should be considered suspect.
                  Reanalyze  sample extracts for these  analytes  after  acceptable
                  calibration is restored'-.

          10.2.5  Verify  calibration  standards periodically  (at least
                  quarterly), by analyzing a QCS.

     10.3 EXTERNAL STANDARD CALIBRATION PROCEDURE

          10.3.1  Prepare calibration standards as  in  Section 10.2.1,  omitting
                  the use of the internal standard.

          10.3.2  Starting with the standard of lowest  concentration,  analyze
                  each calibration standard according  to  Sect.  11.4 and
                  tabulate response (peak height  or area)  versus  the
                  concentration in the  standard.  The  results can be  used  to
                  prepare a  calibration curve  for each  compound.
                  Alternatively, if the ratio  of  response to concentration
                  (calibration  factor)  is a constant over the working range
                  (20% RSD or less),  linearity through  the origin can be
                  assumed and the  average ratio or  calibration  factor can  be         :
                  used in place of a  calibration  curve.

          10.3.3  The working calibration curve or  calibration  factor must be        .
                  verified on each working day by the  procedures  described in
                  Section 10.2.4.                                               -:, 't ,>,
                                                                                ; ' •'?•*&,
          10.3.4  Verify  calibration  standards'periodically  (at least    .     ,'.; ;V.7,V
                  quarterly), by analyzing a QCS.                              ,'-.  '.'•^";
                                                                                  . •''.'^jf'1
11.  PROCEDURE                                                                ••.:v-'?f*
                                                                              . " * :  V-v* ;
     11.1 EXTRACTION (MANUAL, METHOD)                                                ;: :;

          11.1.1  Mark the water meniscus on the  side  of  the sample bottle.for
                  later determination of sample volume  (Sect. 11.1.6).   Add
                  preservative  (Section 8) to  LRBs  and  LFBs.  Fortify the

                                     507-16 !

-------
          .  .sample with 50 /il_ of the  surrogate  standard  solution.   Pour
            the entire sample into  a  2-L  separatory  funnel.

     11.1.2 Adjust the. sample to pH 7 by  adding  50 ml of phosphate
            buffer.  Check pH.  Add acid  or base  if  necessary to obtain
      .  ..• . PH'7. ........

    .,11.1.3 Add 100 g NaCl to the.-sample, seal,  and  shake  to dissolve
          .:  salt*.             '   ...                 .-   ..  .

     11.1.4 Add 60 ml methylene chloride  to the  sample bottle,  seal, and
            shake 30 s to rinse the inner walls.  Transfer the  solvent to
            the separatory funnel and extract the sample by vigorously
            shaking the funnel for 2 min  with periodic venting  to release
            excess pressure.  Allow the organic  layer to separate from
            the water phase for a minimum of 10 min.  If the emulsion
            interface between layers  is more than one third the volume of
            the solvent layer, the analyst must  employ mechanical
            techniques to complete the phase separation. ••- The optimum
            technique depends upon the sample, but may, include  stirring,
         .filtration of the emulsion through glass wool,
            centrifugation, or other physical methods.   Collect the
            methylene chloride extract in a 500-mL Erlenmeyer flask.

     11.1.5 Add a second 60-ml volume
                          of methylene chloride to the sample
uwwo.t ui.u icpcuu UMC cAui'action procedure a second tin
combining the extracts in the Erlenmeyer
            bottle
econd 60-ml volume of methylene chloride
and repeat the extraction procedure a se
ng the extracts in the Erlenmeyer flask;.
            third extraction in the same manner
second time,'
    Perform;a
     11.1.6 Determine the original sample volume by refilling the sample
            bottle to the mark and transferring the water to a 1000-mL
            graduated cylinder.  Record the sample volume to the nearest
11.2 EXTRACTION (AUTOMATED METHOD) — Data presented in this method were
     generated using the automated extraction procedure with the
     mechanical tumbler.          •

    ,,11.2.1 Mark the water meniscus on the side of the sample bottle for
        :  • later determination of sample volume (Sect. 11.2.6).  Add
            preservative to LRBs and LFBs.  Fortify the sample with 50 ill
            of the surrogate standard solution.  If the mechanical
            separatory funnel  shaker is used, pour the entire sample into
  ,          a 2-L separatory funnel.  If the mechanical-tumbler is used,
            pour the entire sample into a tumbler bottle.

     11.2.2 Adjust the sample  to pH 7 by adding 50 ml of phosphate
            buffer.   Check pH.   Add acid or base if necessary to obtain
            pH 7.

     11.2.3 Add 100  g NaCl  to  the sample, seal, and shake  to dissolve
            salt.                                   ,,

     11.2.4 Add 3.0.0  mL methylene chloride to the sample bottle,  seal, and
            shake 30 s to rinse the inner walls.  Transfer the solvent to
            the sample contained in the separatory funnel  or tumbler
                               507-17

-------
            bottle,  seal,  and shake for 10 s,  venting periodically.
            Repeat shaking and venting until  pressure release  is  not
            observed.   Reseal and place sample container in  appropriate
            mechanical  mixing device (separatory funnel  shaker or
            tumbler).   Shake or tumble the sample for 1  hour.   Complete
            mixing of the  organic and aqueous  phases should  be observed
            within about 2 min after starting  the mixing device.

     11.2.5 Remove the sample container from the mixing  device.   If  the
            tumbler is used, pour contents of  tumbler bottle into a  2-L
            separatory funnel.  Allow the organic layer  to separate  from
            the water phase for a minimum of 10 min.  If the emulsion
            interface between layers is more than one third  the volume of
            the solvent layer, the analyst must employ mechanical
            techniques to  complete the phase separation.  The  optimum
            technique depends upon the sample, but may include stirring,
            filtration through glass wool, centrifugation, or  other
            physical methods.  Collect the methylene chloride  extract in
            a 500-mL Erlenmeyer flask...

     11.2.6 Determine the  original sample volume by refilling  the sample
            bottle to the  mark and transferring the water to a 1000-mL
            graduated cylinder.  Record the sample volume to the nearest
            5 ml.

11.3 EXTRACT CONCENTRATION

     11.3.1 Assemble a K-D concentrator by attaching- a 25JmL concentrator
            tube to a 500-mL evaporative flask.  Other concentration
            devices or techniques may be used  in place of the  K-D if the
            requirements of Sect. 9.3 are met.

     11.3.2 Dry the extract by pouring it through a solvent-rinsed drying
            column containing about 10 cm of anhydrous sodium sulfate.
            Collect the extract in the K-D concentrator, and rinse the
            column with 20-30 mL methylene chloride. Alternatively,  add
            about 5 g anhydrous sodium sulfate to the extract  in the
            Erlenmeyer flask; swirl flask to dry extract and allow to sit
            for 15 min. Decant the methylene chloride extract  into the
            K-D concentrator.  Rinse the remaining sodium sulfate with
            two 25-mL portions of methylene chloride and decant the
            rinses into the K-D concentrator.

     11.3.3 Add 1 to 2 clean boiling stones to the evaporative flask and
            attach a macro Snyder column.  Prewet the Snyder column  by
            adding about 1 mL methylene chloride to the top.  Place  the
            K-D apparatus on a hot water bath, 65 to 70°C, so that the
            concentrator tube is partially immersed in the hot water, and
            the entire lower rounded surface of the flask is bathed  with
            hot vapor.  Adjust the vertical position of the apparatus and
            the water temperature as required to complete the
            concentration in 15 to 20 min.  At the proper rate of
            distillation,  the balls of the column will actively chatter,
            but the chambers will not flood.  When the apparent volume of
            liquid reaches 2 mL, remove the K-D apparatus and allow it to
            drain and cool for at least 10 min.

                               507-18

-------
      11.3.4 Remove the Snyder column and rinse the. flask and its lower
             joint into the concentrator tube with 1 to 2 ml of MTBE.   Add
             5-10 mL of MTBE and a fresh boiling stone.  Attach a
             micro-Snyder column to the concentrator tube and prewet the
             column by adding about 0.5 ml of MTBE to the top.  Place  the
             micro K-D apparatus on the water bath so that the
             concentrator tube is partially immersed in the hot water.
             Adjust the vertical position of the apparatus and the water
             temperature as required to complete concentration in 5 to  10
             min.   When the apparent volume of liquid reaches 2 ml,  remove
             the micro K-D from the bath and allow it to drain and cool.
          .   Add 5-10 ml MTBE to the micro K-D and reconcentrate to 2 ml
             Remove the micro K-D from the bath and allow it to drain and
             cool.   Remove the micro Snyder column,  and rinse the walls  of
             the concentrator tube while adjusting the volume to 5.0 ml
             with  MTBE.   NOTE:   If methylene chloride is not completely
             removed  from the final  extract,  it may cause detector
             problems.   If the internal  standard calibration procedure  is
             used,  add 50 /iL  of the internal  standard solution to the
             sample extract,  seal,  and shake to distribute the internal
             standard.

      11.3.5  Transfer extract to an  appropriate- sized TFE-fluorocarbon-
             sealed screw-cap vial  and store,  refrigerated at 4°C,  until
             analysis by GC-NPD.

11.4  GAS CHROMATOGRAPHY

      11.4.1  Sect.  6.8  summarizes  the  recommended  operating  conditions for
             the gas  chromatograph.   Included  in Table  1  are  retention
             times  observed using  this method.   Other  GC  columns  or
             chromatographic  conditions  may  be  used  if  the  requirements of
             Sect.  9  are  met.

      11.4.2  Verify the  calibration  the  system daily  as described  in Sect.
             10.2.4 or  10.3.3.   The  standards and extracts must  be in
             MTBE.

      11.4.3  Inject 2 /il_  of the  sample extract.  Record the resulting peak
             size in  area units.

      11.4.4  If the response  for the peak exceeds the working range of the
             system, dilute the extract and reanalyze.  If using  IS
            calibration, add an appropriate amount of  IS so that the
            extract concentration will match the calibration standards.

11.5 IDENTIFICATION OF ANALYTES

     11.5.1  Identify a sample component by comparison of its retention
            time to the retention time of a reference chromatogram.  If
            the retention time of an unknown compound corresponds, within
            limits, to the retention time of a standard compound, then
            identification is considered positive.

     11.5.2 The width of the retention time window used to make
            identifications should be based upon measurements of actual

                               507-19

-------
                 retention time variations of standards over the course of a
                 day.  Three times the standard deviation of a retention time
                 can be used to calculate a suggested window size for a
                 compound.  However, the experience of the analyst should
                 weigh heavily in the interpretation of chromatograms.

          11.5.3 Identification requires expert judgement when sample
                 components are not resolved chromatographically.  When peaks
                 obviously represent more than one sample component (i.e.,
                 broadened peak with shoulder(s) or valley between two or more
                 maxima), or any time doubt exists over the identification of
                 a peak on a chromatogram, appropriate alternative techniques
                 to help confirm peak identification, need be employed.  For
                 example, more positive identification may be made by the use
                 of an alternative detector which operates on a
                 chemical/physical principle different from that originally
                 used, e.g., mass spectrometry (6), or the use of a second
                 chromatography column.  A suggested alternative column is
                 described in Sect. 6.8.

12.  CALCULATIONS

     12.1  Calculate analyte concentrations in the sample from the response
           for the analyte using the procedure for multi-point calibration
           described in Sect. 10.  Do not use the daily calibration
           verification standard to calculate analye amounts in samples.

     12.2  If the internal standard calibration procedure is used, calculate
           the concentration (C) in the sample using the response factor  (RF)
           determined in Sect. 10.2.2 and Equation 2, or determine sample
           concentration from the calibration curve.

                    C (pg/L) =      (AS)(U        Equation 2

                               (A,s)(RF)(Vo)

           where:

           As    -  Response for the parameter to be measured.

           Ais    -  Response for the internal standard.

           I     =  Amount of internal standard added to each extract  (Mg).

           Vo    =  Volume of water extracted (L).

     12.3  If the external standard calibration procedure is used, calculate
           the amount of material injected from the peak response using the
           calibration curve or calibration factor determined in Sect.  10.3.2.
           The concentration (C) in the sample can be calculated from
           Equation 3.
                   C  (M9/L) =   v M t;           Equation 3

                                (V,)(V8)



                                    507-20

-------
         ,   where:

            A     =  Amount of material  injected (ng).

            V,.     =  Volume of extract injected

            Vt     =  Volume of total  extract

            Vs     =  Volume of water  extracted (ml).


 13.   PRECISION AND ACCURACY

      13.1   In  a single laboratory, analyte recoveries  from reagent water were
            used to determine analyte MDLs, EDLs (Table 3)  and  demonstrate
            method  range.   Analytes were  divided into five  groups  for recovery
            studies.   Analyte recoveries  and standard deviation about the
           . percent recoveries at  one concentration  are given  in Table 2.

      13.2   In  a single laboratory, analyte recoveries  from two standard
            synthetic  ground waters were  determined  at  one  concentration  level.
            Results were used to demonstrate applicability  of the  method  to
            different  ground water matrices.   Analyte recoveries from the  two
            synthetic  matrices are given  in Table  2.

 14.   POLLUTION PREVENTION

      14.1   This  method uses significant  volumes of organic  solvents.   It  is
            highly  recommended that laboratories use  solvent recovery systems
            to  recover used  solvent as sample  extracts  are  being concentrated.
            Recovered  solvents should  be recycled  or  properly disposed  of.

      14.2   For information  about  pollution  prevention  that may  be  applicable
            to  laboratory operations,  consult  "Less is  Better:   Laboratory
            Chemical Management for Waste Reduction"  available  from  the
            American Chemical  Society's Department of Government Relations and
            Science Policy,  1155 16th  Street N.W., Washington, D.C.  20036.

 15.  WASTE  MANAGEMENT

     15.1   It  is the  laboratory's responsibility to comply with all  federal,
            state,  and  local  regulations governing waste management,  particu-
            larly the  hazardous waste  identification rules and land disposal
            restrictions.  The laboratory using this method has  the responsi-
            bility to  protect  the  air, water, and land by minimizing  and
            controlling  all  releases from fume hoods and bench operations.
            Compliance  is also required with any sewage discharge permits and
            regulations.  For  further information on waste management, consult
            "The Waste Management Manual  for Laboratory Personnel," also
           available from the American Chemical Society at the address in
           Sect. 14.2.

16.  REFERENCES
     	                                    •          ,      ;

     1.   ASTM Annual Book of Standards,  Part 11, Volume 11.02,  D3694-82,
         "Standard Practice for Preparation of Sample Containers and for

                                    507-21

-------
   "Preservation," American Society for Testing and Materials, Philadel-
    phia, PA, 1986.

2.  "Carcinogens - Working with Carcinogens," Department of Health,
    Education, and Welfare, Public Health Service, Center for Disease
    Control, National Institute for Occupational Safety and Health,
    Publication No. 77-206, Aug. 1977.

3.  "OSHA Safety and Health Standards, General Industry," (29 CFR 1910),
    Occupational Safety and Health Administration, OSHA 2206, (Revised,
    January 1976).

4.  "Safety in Academic Chemistry Laboratories," American Chemical
    Society Publication, Committee on Chemical Safety, 3rd Edition, ,1979.

5.  ASTM Annual Book of Standards, Part 11, Volume 11.01, D3370-82,
    "Standard Practice for Sampling Water," American Society for Testing
    and Materials, Philadelphia, PA, 1986.

6.  Munch, J. W., "Method 525.2-Determination of Organic Compounds in
    Drinking Water by Liquid-Solid Extraction and Capillary Column
    Chromatography/ Mass Spectrometry" in Methods for the Determination
    of Organic Compounds in Drinking Water: Supplement 3  (1995).  USEPA,
    National Exposure Research Laboratory, Cincinnati, Ohio 45268.
                                507-22

-------
                 TABLE 1.   RETENTION TIMES FOR METHOD ANALYTES

                                                      Retention Time3
Analyte                                               Col. 1    Col. 2
l,3-Dimethyl-2-nitrobenzene( surrogate)
Dichlorvos
Disulfoton sulfoxide
EPIC
Butyl ate
Mevinphos
Vernolate
Pebulate
Tebuthiuron
Molinate
Ethoprop
Cycloate
Chlorpropham
Atraton
Simazine
Prometon
Atrazine
Propazine
Terbufos
Pronamide
Diazinon
Disulfoton
Terbacil
Metribuzin
Methyl paraoxon
Simetryn
Alachlor
Ametryn
Prometryn
Terbutryn
Bromacil
Metolachlor
Triademefon
MGK 264 (c)
Diphenamid
Stirofos
Disulfoton sulfone
Butachlor
Fenamiphos
Napropamide
Tricyclazole
Merphos (d)
Carboxin
Norflurazon
Triphenyl phosphate (int. std.)
14.48
16.54
19.08
20.07
22.47
22.51
22.94
23.41
25.15
25.66
28.58
28.58
29.09
31.26
31.49
31.58
31.77
32.01
32.57
32.76
33.23
33.42
33.79
35.20
35.58
35.72
35.96
36.00
36.14
36.80
37.22
37.74
38.12
38,73
38.87
41.27
41.31
41.45
41.78
41.83
42.25
42.35
42.77
45.92
47
(b)
15.35
(b)
16.57
18.47
21.92
19.25
19.73
42.77
22.47
26.42
29.67
(b)
29.97
31.32
30.00
31.23
31.13
(b)
32.63
(b)
30.90
(b)
34.73
34.10
34.55
34.10
34.52
34.23
34.80
40.00
35.70
37.00
36.73
37.97
39.65
42.42
39.00
41.00
(b)
44.33
39.28
42.05
47.58
45.40
                                    507-23

-------
                             TABLE 1  (CONTINUED)


                                                    Retention Time3.
Analyte                                              Col.l     Col.2
Hexazinone
Fenarimol
Fl undone
46.58
51.32
56.68
47.80
50.02
59.07
8  Columns and analytical  conditions are described in Sect.  6.8.1  and 6.8.2,

b  Data not available

c  MGK 264 gives two peaks; peak identified in this table used for
   quantification.

   Merphos is converted to S,S,S-tributylphosphoro-trithioate (DEF)  in the
   hot GC injection port;  DEF is actually detected using these analyses
   conditions.
d
                                     507-24

-------
TABLE 2.
SINGLE LABORATORY ACCURACY AND PRECISION FOR ANALYTES FROM
REAGENT WATER AND SYNTHETIC GROUNDWATERS3
Analyte
A achlor
Ametryn
Atraton
Atrazine
Bromacil
Butachlo
Butyl ate
Carboxin
Chlorpropham
Cycloate
Diazinon
Dichlorvos
Diphenamid
Disulfoton
Disulfoton sulfone
Disulfoton sulfoxide
EPIC
Ethoprop
Fenamiphos
Fenarimol
Fluridone
Hexazinone
Merphos
Methyl paraoxon
Metol achlor
Metribuzin
Mevinphos
MGK 264
Mol inate
Napropamide
Norflurazon
Pebulate
Prometon
Prometryn
Pronamide
Propazine
Simazine
Simetryn
Stirofos
Tebuthiuron
Terbacil
Terbufos
Terbutryn
• Cone.
378
• 20
6
1.3
25
3.8 •
1.5.
6
5
2.5
2.5
25
6
.3
, 7.5
3.8
2.5
1.9
' 10
3.8
38
7.6
2.5
25
7.5
1.5
50
5
1.5
2.5
5.
1.3
3
1.9
7.6
1.3
0.75
2.5
7.6
13
45
5
2.5
Reagent
Water
Rb SRC
95
91
91
92
91
96
97
102
93
89
115
97
93
89
98
87
85
103
90
99
87
90
96
98
93
101
95
100
98
101
94
94
78
93
91
92
100
99
98
84
97
97
94
11
10
11
8
9
4
21
4
11
9
7 .:
6
8
10
10
11
9
5
8
5
9
7
8
10
4
5
11
4
18
6
5
9
9
8
10
8
7
5
6
9
6
4
9
Synthetic
Water ld
R SR
82
102 .
84
89
81
93
36
98
82
97
83
86
88
107
92
88
83
91
87
89
91
86
90
97
92
99
93
91
83
89
.101
80
89
91
84
89
86
88
84
85
86
80
91
6
11
7
6
5
15
8
13
7
14
8
6
4
12
5
22
5
7
5
6
11
6 '
4
8
10
10
6
11
8
5
15
6
5
8
7
6
5
4
6
10
5
6
8
Synthetic
Water 2e
R SR
90
96
91
92
88
84
83
87
93
93 .
84
106
93
95
! 96
54
86
79
89
89
86
95
92
94
84
86
92
83
89
104
87
98
63
93
92
92
103
103
95
98,
102
77
92
8
4
8
5
8
5
8
5
8
3
3
16
5
5
3
19
4
3
2
6
10
g
4
4
4
4
4
6
9
18
4
15
2
4
8
5
14
14
10
13
12
7
4
                                    507-25

-------
                             TABLE 2.   (CONTINUED)
Analyte
Cone.
M9/L
Reagent
Water
Rb SRC
Synthetic
Water ld
R SR
Synthetic
Water 2e
R SR
Triademefon
Tricyclazole
Vernolate
 6.5
10
 1.3
93
86
93
8
7
6
94
90
79
5
6
9
95
90
81
 5
11
 2
Data corrected for blank and represent the analysis of 7-8 samples using
mechanical tumbling and internal standard calibration.

R = average percent recovery.

S = standard deviation of the percent recovery.

Corrected for amount found in blank; Absopure Nature Artesian Spring Water
Obtained from the Absopure Water Company in Plymouth, Michigan.

Corrected for amount found in blank; reagent water fortified with fulvic acid
at the 1 mg/L concentration  level.  A well-characterized fulvic acid,
available from the International Humic Substances Society  (associated with
the United States Geological Survey in Denver, Colorado), was used.
                                  507-26

-------
TABLE 3.
SINGLE LABORATORY ACCURACY, PRECISION, METHOD DETECTION LIMITS
(MDLs) AND ESTIMATED DETECTION LIMITS (EDLs) FOR ANALYTES FROM
REAGENT WATER
Analyte
Fortified Cone.    Na   Recovery   RSD
                                                        MDL
EDLC
Alachlor
Ametryn
Atraton
Atrazine
Bromacil
Butachlor
Butyl ate
Carboxin
Chlorpropham
Cycloate
Diazinon
Dichlorvos
Diphenamid
Disulfoton
Disulfoton sulfone
Disulfoton sulfoxide
EPIC
Ethoprop
Fenamiphos
Fenarimol
Fluridone
Hexazinone
Merphos
Methyl paraoxon
Metolachlor
Metribuzin
Mevinphos
MGK 264
Molinate
Napropamide
Norflurazon
Pebulate
Prometon
Prometryn
Pronamide
Propazine
Simazine
Simetryn
Stirofos
Tebuthiuron
Terbacil
Terbufos
Terbutryn
0.38
2.0
0.60
0.13
2.5
0.38
0.15
0.60
0.50
0.25
0.25
2.5
0.60
0.30
3.8
0.38
0.25
0.19
1.0
0.38
3.8
0.76
0.25
2.5
0.75
0.15
5.0
0.50
0.15
0.25
0.50
0.13
0.30
0.19
0.76
0.13
0.075
0.25
0..76
1.3
4.5
0.5
0.25
8
8
8
8
8
8
8
8
8
8
8
8
8
8
7
7
8
8
8
8
7
8
8
8
8
8
8
8
8
8
8
8
7
8
8
8
8
8
8
8
8
8
8
119
100
120
101
113
99
93
101
124
101
94
78
84
100
94
110
87
108
91
92
78
127
101
100
94
114
92
101
117
97
86
84
48
88
123
93
99
97
121
101
100
91
91
10
3
8
4
8
11
13
10
11
3
18
5
5
3
6
6
12
3
4
19
30
5
5
4
9 •
6
6
12
12
9
8
7
9
5
10
4
6
5
7
15
4
4
4
0.14
0.20
0.17
0.015
0.69
0.12
0.053
0.18
0.20
0.022
0.13
0.28
0.082
0.029
0.63
0.082
0.080
0.021
0.12
0.20
2.8
0.15 •
0.040
0.30
0.19
0.029
0.87
0.19
0.061
0.069
0.098
0.022
0.041
0.024
0.28
0.014
0.014
0.035
0.18
0.58
0.56
0.054
0.031
0.38
2.0
0.6
0.13
2.5
0.38
0.15
0.60
0.50
0.25
0.25
2.5
0.60
0.30
3.8,
0.38
0.25
0.19
1.0
0.38
3.8
0.76
0.25
2.5
0.75
0.15
5.0
0.50
0.15
0.25
0.50
0.13
0.30
0.19
0.76
0.13
0.075
0.25
0.76
1.3
4.5
0.50
0.25
                                   507-27

-------
                      TABLE 3.   (CONTINUED)
MDL Cone.
Analyte
Tri ademef on
Tricyclazole
Vernol ate
MQ/L Na
0.65 8
1.0 8
0.13 8
. Recovery
#g/L
95
216
100
RSD
%
5
3
14
MDLb
W/L
0.093
0.21
0.055
EDLC
yg/L
0.65
1.0
0.13
* N - Number of Replicates
b With thi
=><;e data, the method c
detection 1
imits (MDL) in
the table
s were
calculated using the formula:

    MDL - S t(n_., j.atpha = 0.99)

    where:

    t  , „  , u   n oo^ =  Student's t value for the 99% confidence level
     
-------
                                                                                  c
                                                                                  cu
                CU
                cu
                i_
=>

o
(/}
 0
 UJ


 al
            U E

            o o»
                          CO

                           A
                          to
                           CU
                           £=
                           03
                          C
                          O
                          CU
                          cu
                         a
                         in
                         O
                                  TO
                                 o
                                 CNJ
                                 V
                               a.

                               v
                                 o
                                 00


                                O

                                ir>
                                      1^.

                                      o

                                      A
 O
 00
 cu
a;
                                     o 10
                                     CO i— i
                                       o o
                                                                                  nj
                                                                                  00
                                                                                  o
                                                                                  cu
                                                                                  oo
                                          73

                                          S
                                                                                cu
                                                                               'a.
                                                                                 OJ
                                                                                          ro
                                                                                          CU
                                                                                          a.

                                                                                          cu
                                                                                          Ol
                                                                                          ro

                                                                                          CU

                                                                                          re

                                                                                          CU
cu
+J
03
t—
O
sz
&_
CU
>

r"-
•r-*
0
03
E
0
s_
en

c cu
o c
•4-> ••—
CU N
E 03
O i~

Q. rf
Ul
00
     i-
•r-     CJ1
 >     o
••—    +j
-M     03
••-     E
 00     o
 c:     i_
 cu    ^=
CO    l_5
                              I

                               ai     CD
                               Q.     0

                               
                                                                                         a>
                                                                               r—        "O
                                                                                          00
                                                                                          03
                                                                                          03
                                                                                          cu
                                                                                          CL

                                                                                          O
                                                                                         CU
                                                  C
                                                  
                                                                                       U
                                                                                       c  cu
                                                                                       CU -C
                                                                                       S- -M
                                                                                       cu

                                                                                      H-'S
                                                                          CU
                                                                      cu  c
                                                                      JZ -i-
                                                                      •*-> 1 —
                                                                          cu
                                                                      00  00
                                                                      •i-  OS
                                                                                  CD
                                                                                         o
                                                                                         00
                                                                                         cu
                                                                                        o:
                                                                                                            cu ,

                                                                                                            cu
                                                                                                                 CU
                                                                                                                03
                                                           507-29

-------
THIS PAGE LEFT BLANK INTENTIONALLY
              507-30

-------
  METHOD 508. DETERMINATION OF CHLORINATED PESTICIDES IN WATER BY GAS
              CHROMATOGRAPHY WITH AN ELECTRON CAPTURE DETECTOR
                              Revision 3.1
                      Edited by J.W.  Munch  (1995)
J. J. Lichtenberg, J. E. Longbottom, T. A. Bellar, J. W. Eichelberger
   and R.  C.  Dressman -  EPA 600/4-81-053,  Revision 1.0 (1981)


D. J. Munch  (USEPA,  Office of Water)  and  T.  Engel  (Battelle Columbus
Laboratories) - National Pesticide Survey Method 2, Revision 2.0 (1987)

R. L. Graves -  Method 508,  Revision 3.0  (1989)
                 NATIONAL EXPOSURE RESEARCH LABORATORY
                  OFFICE OF RESEARCH AND DEVELOPMENT
                 U.S. ENVIRONMENTAL PROTECTION AGENCY
                        CINCINNATI, OHIO 45268
                                 508-1

-------
                                  METHOD 508

             DETERMINATION  OF  CHLORINATED  PESTICIDES  IN WATER  BY
             GAS CHROHATOGRAPHY WITH AN ELECTRON CAPTURE DETECTOR
1.   SCOPE AND APPLICATION

     1.1  This is a gas chromatographic (GC) method applicable to the
          determination of certain chlorinated pesticides in groundwater and
          finished drinking water.  The following compounds can be determined
          using this method:
               Analvte

               Aldrin
               Chlordane-alpha
               Chlordane-gamma
               Chlorneb
               Chlorobenzilate(a)
               Chlorothalonil
               DC PA
               4,4'-DDD
               4,4'-DDE
               4,4'-DDT
               Dieldrin
               Endosulfan  I
               Endosulfan  II
               Endosulfan  sulfate
               Endrin
               Endrin  aldehyde
               Etridiazole
               HCH-alpha
               HCH-beta
               HCH-delta(a)
               HCH-gamma  (Lindane)
               Heptachlor
               Heptachlor  epoxide
               Hexachlorobenzene
               Methoxychlor
               cis-Permethrin
               trans-Permethrin
               Propachlor
               Trifluralin
               Aroclor 1016*
               Aroclor 1221*
               Aroclor 1232*
               Aroclor 1242*
               Aroclor 1248*
               Aroclor 1254*
               Aroclor 1260*
Chemical Abstract Service
    Registry Number '
         309-
        5103-
        5103-
        2675-
         501-
        2921-
        1861-
          72-
          72-
          50-
          60-
         959-
       33213-
        1031-
          72-
        7421-
        2593-
         319-
         319-
         319-
          58
          76-
        1024
         118
          72
       52645
       52645
        1918
        1582
       12674
       11104
       11141
       53469
       12672
       11097
       11096
00-2
71-9
74-2
77-6
15-6
88-2
32-1
54-8
55-9
29-3
57-1
98-8
65-9
07-8
20-8
93-4
•15-9
•84-6
•85-7'
•86-8
•89-9
•44-8
-57-3
-74-1
-43-5
-53-1
-53-1
-16-7
-09-8
-11-2
-28-2
-16-5
-21-9
-29-6
-69-1
-82-5
                                      508-2

-------
                Toxaphene*                       8001-35-2
                Chlordane*                         57-74-9

           M Jh! !£JracJ1on conditi°ns of this method are comparable to USEPA
           Method 608, which does measure the multicomponent constituents-
           commercial polychlorinated biphenyl (PCS) mixtures (Aroclors)
           toxaphene, and chlordane.  The extract derived from this procedure
           may be analyzed for these constituents by using the GC conditions
           prescribed in either Method 608 (packed column) or Methods 505
           SOS.ror 525.2 (capillary column)(l).   The columns used -in this
           method may well be adequate,  however,  no data were collected for
           these constituents during methods development.

           (a)   Chlorbenzilate and HCH-delta are  only qualitatively identified
           and  are not quantitated because control  over precision has not been
           accomplished.

      1.2  This method has been validated  in a single laboratory and estimated
           detection limits  (EDLs) and method detection limits  (MDLs)  have been
           determined for the analytes above (Sect.  13).   Observed detection  '
           limits  may vary between waters,  depending  upon the nature of
                         ^  the sample matrix and  the specific  instrumentation
1.3  This method is restricted to use by or under the supervision of
     analysts experienced in the use of GC and in the interpretation of
     gas chromatograms.  Each analyst must demonstrate the ability to
1.4
                                                                         o
          generate  acceptable  results with  this method  using the procedure
          described  in  Sect. 9.3.                             ,

          Analytes  that  are not  separated chromatographically, i.e., analytes
          which have very  similar retention times, cannot be individually
          identified and measured in the same calibration mixture or water
          sample unless  an alternative technique for identification and
          quantitation exist, (Sect. 11.5).

     1.5  When this method is  used to analyze unfamiliar samples for any or
          all of the analytes  above, analyte identifications must be confirmed
          by at least one additional qualitative technique.

2.   SUMMARY OF METHOD

     2.1  A measured volume of sample of approximately 1 L is solvent
          extracted with methylene chloride by shaking in a  separatory funnel
          or mechanical  tumbling in a bottle.   The methylene chloride extract
          is isolated,  dried  and concentrated  to a volume of 5  mL after
          solvent  substitution  with methyl  tert-butyl  ether  (MTBE) .   Chroma-
          tographic conditions  are described which permit the separation  and
          measurement of the  analytes  in  the extract  by  capillary column  GC
          with  an  electron  capture detector  (ECO).
                                    508-3

-------
3.   DEFINITIONS
     3.1
     3.2
     3.3
      3.4
      3.5
      3.6
      3.7
      3.8
INTERNAL STANDARD — A pure analyte(s) added to a solution in known
amount(s) and used to measure the relative responses of other method
analytes and surrogates that are components of the same solution.
The internal standard must be an analyte that is not a sample
component.

SURROGATE ANALYTE — A pure analyte(s), which is extremely unlikely
to be found in any sample, and which is added to a sample aliquot in
known amount(s) before extraction and is measured with the same
procedures used to measure other sample components.  The purpose of
a surrogate analyte is to monitor method performance with each
sample.

LABORATORY DUPLICATES  (LD1 and LD2) — Two sample aliquots taken in
the analytical laboratory and analyzed separately with identical
procedures.  Analyses  of LD1 and LD2 give  a measure of the precision
associated with laboratory procedures, but .not with sample
collection, preservation, or storage procedures.

FIELD DUPLICATES  (FD1  and FD2) — Two  separate samples collected at
the same  time  and place under identical circumstances  and treated
exactly  the same  throughout field and  laboratory  procedures.
Analyses  of FD1 and  FD2 give a measure of  the  precision  associated
with sample collection, preservation  and  storage,  as well as with
laboratory  procedures.

LABORATORY  REAGENT  BLANK  (LRB) — An  aliquot  of  reagent  water  that
is treated  exactly  as  a sample  including  exposure to  all  glassware,
equipment,  solvents,  reagents,  internal standards,  and surrogates
that are used  with  other  samples.   The LRB is  used  to  determine  if
method  analytes or  other  interferences are present  in  the laboratory
environment,  the  reagents,  or the  apparatus.

FIELD  REAGENT  BLANK (FRB)  —  Reagent  water placed in  a sample
container in  the  laboratory and  treated as a  sample in all  respects,
including exposure  to sampling  site conditions,  storage,
preservation  and  all  analytical  procedures.   The purpose of the  FRB
is to  determine  if  method analytes  or other interferences are
present in the field environment.

LABORATORY PERFORMANCE CHECK SOLUTION (LPC)  — A solution of method
analytes, surrogate compounds,  and internal  standards used  to
evaluate the performance  of the instrument system with respect to a
defined set of method criteria.

LABORATORY FORTIFIED BLANK (LFB) — An aliquot of reagent water to
which  known quantities of the method analytes are added in  the
laboratory.  The LFB is analyzed exactly  like a sample, and its
 purpose is to determine whether the methodology is in control, and
                                      508-4

-------
          whether  the  laboratory  is  capable  of making  accurate  and precise
          measurements  at  the  required method detection  limit.

     3.9  LABORATORY FORTIFIED SAMPLE MATRIX (LFM) —  An  aliquot of  an
          environmental  sample to which  known quantities  of the method
        .  analytes  are  added in the  laboratory.  The LFM  is analyzed exactly
          like  a sample, and its purpose  is  to determine  whether the sample
          matrix contributes bias to the  analytical results.  The background
          concentrations of the analytes  in  the sample matrix must be
          determined in  a  separate aliquot and the measured values in the LFM
          corrected for  background concentrations.

     3.10 STOCK STANDARD SOLUTION — A concentrated solution containing a
          single certified standard  that  is  a method analyte, or a
          concentrated  solution of a single  analyte prepared in the laboratory
          with  an  assayed  reference  compound.  Stock standard solutions are
          used  to  prepare  primary dilution standards.

     3.11 PRIMARY  DILUTION STANDARD  SOLUTION — A solution of several analytes
          prepared  in the  laboratory from stock standard  solutions and diluted
          as needed to  prepare calibration solutions and  other needed analyte
          solutions.

     3.12 CALIBRATION STANDARD  (CAL) — A solution prepared from the primary
          dilution standard solution and stock standard solutions of the
          internal standards and surrogate analytes.  The CAL solutions are
          used  to calibrate the instrument response with  respect to analyte
          concentration.

     3.13 QUALITY CONTROL SAMPLE (QCS) — a  sample matrix containing method
          analytes or a solution of method analytes in a water miscible
          solvent which is used to fortify reagent water  or environmental
          samples.  The QCS is obtained from a source external  to the
          laboratory,  and is used to check laboratory performance with
          externally prepared test materials.

4.   INTERFERENCES

     4.1  Method interferences may be caused by contaminants in solvents,,
          reagents, glassware and other sample processing apparatus that lead
          to discrete artifacts or elevated baselines in gas chromatograms.
          All reagents  and apparatus must be routinely demonstrated to be  free
          from  interferences under the conditions of the analysis by running
          laboratory reagent blanks as described in Sect. 9.2.

          4.1.1  Glassware must-be scrupulously cleaned (2).   Cle.an all  glass-
                 ware as soon as possible after use by thoroughly rinsing  with
                 the last solvent used in it.  Follow by washing with hot
                 water  and detergent and thorough rinsing with  tap and reagent
                 water.  Drain  dry,  and heat in an oven or muffle furnace  at
                 400°C  for 1 hour.   Do not heat volumetric glassware.
                 Thermally stable materials such as PCBs might  not be

                                     508-5

-------
            eliminated by this treatment.  Thorough rinsing with acetone
            may be substituted for the heating.   After drying and
            cooling, seal and store glassware in a clean environment to
            prevent any accumulation of dust or other contaminants.
            Store inverted or capped with aluminum foil.

     4.1.2  The use of high purity reagents and solvents helps to
            minimize interference problems.  Purification of solvents by
            distillation in all-glass systems may be required.  WARNING:
            When a solvent is purified, stabilizers added by,the
            manufacturer are removed thus potentially making the solvent
            hazardous.  Also, when a solvent is purified, preservatives
            added by the manufacturer are removed thus potentially
            reducing  the shelf-life.

4.2  Interferences by phthalate esters can pose a major problem in pesti-
     cide analysis when using the electron capture detector.  These
     compounds generally appear in the chromatogram as large peaks.
     Common flexible plastics contain varying amounts of phthalates  that
     are easily extracted or leached during laboratory operations.  Cross
     contamination of clean glassware routinely occurs when plastics are
     handled during extraction steps, especially when solvent-wetted
     surfaces are handled.  Interferences from phthalates can best be
     minimized by avoiding the use of plastics in the laboratory.
     Exhaustive cleanup of reagents and glassware may be required to
     eliminate background phthalate contamination.    '   :.

4.3  Interfering contamination may occur when a sample containing Tow
     concentrations of analytes is analyzed immediately following a
     sample containing relatively high concentrations of analytes.
     Between-sample rinsing of the sample syringe and associated
     equipment with MTBE can minimize sample cross contamination.  After
     analysis of a sample containing high concentrations of analytes, one
     or more injections of MTBE should be made to ensure that accurate
     values are obtained for the next sample.

4.4  Matrix interferences may be caused by contaminants that are
     coextracted from the sample.  Also, note that all the analytes
     listed in the Scope and Application Section are not resolved from
     each other on any one column, i.e., one analyte of interest may be
     an interferant for another analyte of interest.  The extent of
     matrix interferences will vary considerably from source to source,
     depending upon the water sampled.  Analyte identifications should be
     confirmed (Sect. 11.5).

4.5  It is important that samples and standards be contained in the  same
     solvent, i.e., the solvent for final working standards must be  the
     same as the final solvent used in sample preparation.   If this  is
     not the case, chromatographic comparability of standards to sample
     may be affected.
                                508-6

-------
5.   SAFETY

     5.1  The toxicity  or  carcinogenicity of each  reagent used  in this method
          has not been  precisely defined; however, each chemical compound must
          be treated as  a  potential health hazard.  Accordingly, exposure to
          these chemicals  must be reduced to the lowest possible level.  The
          laboratory is  responsible for maintaining a current awareness file
          of OSHA regulations regarding the safe handling of the chemicals
          specified in  this method.  A reference file of material safety data
          sheets should  also be made available to  all personnel involved in
          the chemical  analysis.  Additional references to laboratory safety
          are available  and have been identified (3-5) for the  information of
          the analyst.

     5.2  WARNING:  When a solvent is purified stabilizers added by the
          manufacturer are removed thus potentially making the  solvent
          hazardous.

6-   EQUIPMENT AND SUPPLIES (All  specifications are suggested.   Catalog
     numbers are included for illustration only.)

     6.1  SAMPLE BOTTLE — Borosil icate,  1-L volume with graduations (Wheaton
          Media/Lab bottle 219820 or equivalent), fitted with screw caps lined
          with TFE-fluorocarbon.   Protect samples from light.  Amber bottles
          may be used.   The container mu.st be washed  and dried  as described  in
          Sect.  4.1.1  before use  to minimize contamination.   Cap liners are
          cut to fit from sheets  (Pierce  Catalog  No.  012736)  and extracted
          with methanol  overnight prior to use.

     6.2  GLASSWARE

          6.2.1   Separatory funnel  —  2000-mL,  with TFE-fluorocarbon stopcock
                 ground  glass  or  TFE-fluorocarbon stopper.

          6.2.2   Tumbler bottle  1.7-L  (Wheaton  Roller  Culture Vessel  or
                 equivalent),  with  TFE-fluorocarbon lined  screw  cap.   Cap
                 liners  are cut to  fit  from  sheets  (Pierce Catalog No.  012736)
                 and extracted with methanol overnight prior  to  use.

          6.2.3   Flask,  Erlenmeyer  — 500-mL.

          6.2.4   Concentrator  tube, Kuderna-Danish  (K-D) 10-  or  25-mL
                graduated  (Kontes K-570050-1025  or K-570050-2525 or
                equivalent).  Calibration must be  checked at the volumes
                employed in the test.  Ground glass stoppers are used to
                prevent  evaporation of extracts.

          6.2.5  Evaporative flask, K-D 500-mL (Kontes K-570001-0500 or
                equivalent).  Attach to concentrator tube with  springs.

          6.2.6  Snyder column, K-D three-ball macro (Kontes K-503000-0121 or
                equivalent).

                                    508-7

-------
     6.2.7  Snyder column, K-D two-ball micro (Kontes K-569001-0219 or
            equivalent).

     6.2.8  Vials — Glass, 5- to 10-mL capacity with TFE-fluorocarbon
            lined screw cap.

6.3  SEPARATORY FUNNEL SHAKER — Capable of holding 2-L separatory
     funnels and shaking them with rocking motion to achieve thorough
     mixing of separatory funnel contents (available from Eberbach Co. in
     Ann Arbor, MI or other suppliers).

6.4  TUMBLER — Capable of holding tumbler bottles and tumbling them
     end-over-end at 30 turns/min (Associated Design and Mfg. Co.,
     Alexandria, VA or other suppliers.).

6.5  BOILING STONES CARBORUNDUM, #12 granules (Arthur H. Thomas Co.
     #1590-033 or equivalent).  Heat at 400°C for 30 min prior to use.
     Cool and store in a desiccator.
6.6
6.7
6.8
WATER BATH — Heated, capable of temperature control (± 2°C).
bath should be used in a hood.
                                                                    The
BALANCE — Analytical, capable of accurately weighing to the. nearest
0.0001 g.

GAS CHROMATOGRAPH — Analytical system complete with temperature
programmable GC suitable for use with capillary columns and all
required accessories including syringes, analytical columns, gases,
detector and stripchart recorder.  A data system is recommended for
measuring peak areas.  Table 1 lists retention times observed for
method analytes using the columns and analytical conditions
described below.

6.8.1  Column 1 (Primary column) — 30 m long x 0.25 mm I.D. DB-5
       bonded fused silica column, 0.25 fim film thickness (J&W
       Scientific).  Helium carrier gas flow is established at 30
       cm/sec linear velocity and oven temperature is programmed
       from 60°C to 300°C at 4°C/min.  Data presented in this method
       were obtained using this column.  The injection volume was 2
       /iL splitless mode with a 45 sec. delay.  The injector
       temperature was 250°C and the detector temperature was 320°C.
       Column performance criteria are presented in Table 4 (See
       Section 9.9).  Alternative columns may be used in accordance
       with the provisions described in Sect. 9.4.

6.8.2  Column 2 (Alternative column) — 30 m long x 0.25 mm
       I.D.DB-1701 bonded fused silica column, 0.25 jitm film
       thickness (J&W Scientific).  Helium carrier gas flow is
       established at 30 cm/sec linear velocity and oven temperature
       is programmed from 60°C to 300°C at 4°C/min.
                                508-8

-------
     6.8.3  Detector  —  Electron  capture.  This detector has proven
            effective in  the  analysis of fortified reagent and artificial
            ground waters.

REAGENTS AND STANDARDS -  - WARNING:  When a solvent  is purified,
stabilizers added by  the  manufacturer are removed thus potentially making
the  solvent hazardous.  Also, when  a solvent  is purified, preservatives
added  by the manufacturer are removed thus potentially reducing the
shelf-life.

7.1  ACETONE, methylene  chloride, MTBE — Distilled-in-glass quality or
     equivalent.

7.2  PHOSPHATE  BUFFER, pH 7  Prepare by mixing  29.6 ml 0.1 N HC1 and 50 ml
     0.1 M dipotassium phosphate.

7.3  SODIUM CHLORIDE,  crystal, ACS  grade.  Heat treat in a shallow tray
     at 400°C for a minimum  of 4  hours to remove  interfering organic
     substances.  Store  in a  glass  bottle (not plastic) to avoid
     phthalate  contamination.

7.4  SODIUM SULFATE,  granular, anhydrous, ACS  grade.  Heat treat  in a
     shallow tray at  450°C for a  minimum of 4  hours  to remove  interfering
     organic substances.  Store in  a glass bottle (not plastic) to avoid
     phthalate  contamination.

7.5  SODIUM THIOSULFATE,  granular,  anhydrous,  ACS grade.        .

7.6  PENTACHLORONITROBENZENE  (PCNB) 98% purity, for  use as internal
     standard.

7.7  DECACHLOROBIPHENYL  (DCB) 96% purity, for  use as surrogate standard
     (available from  Chemicals Procurement Inc.).

7.8  MERCURIC CHLORIDE — ACS grade — for use as a  bactericide
     (optional).

J .9  REAGENT WATER — Reagent water is defined as water that is
     reasonably free  of  contamination that would  prevent the
     determination of any analyte of interest.  Reagent water  used to
     generate the validation  data in this method  was distilled water
     obtained from the Magnetic Springs Water  Co., Columbus, Ohio.
 7.10  STOCK. STANDARD  SOLUTIONS  (1.00  /jg/ML) —  Stock  standard  solutions
      may  be  purchased  as  certified solutions or  prepared  from pure
      standard  materials using  the following procedure:

      7.10.1  Prepare  stock standard solutions by  accurately weighing
             approximately 0.0100 g of pure material .  Dissolve the
             material  in MTBE  and dilute  to volume  in  a  10-mL  volumetric
             flask.   Larger volumes may be used at  the convenience of  the
             analyst.   If  compound purity is certified at  96%  or greater,

                                508-9

-------
             the weight  may  be  used without  correction to  calculate the •
             concentration of the  stock  standard.  Commercially prepared
             stock  standards may be used  at  any concentration  if they are
             certified by the manufacturer or  by an  independent source.

     7.10.2  Transfer the stock standard  solutions into TFE-fluoro-
             carbon-sealed screw cap  amber vials.  Store at room temper-
             ature  and protect  from light.

     7.10.3  Stock  standard  solutions  should be replaced after two months
             or sooner if comparison with laboratory fortified blanks, or
             QC samples  indicate a problem.

7.11 INTERNAL STANDARD  SOLUTION -- Prepare  an internal standard
     fortifying solution by accurately weighing approximately 0.0010 g of
     pure PCNB.  Dissolve the  PCNB in MTBE  and dilute to  volume in a
     10-mL volumetric flask.   Transfer the  internal standard  solution to
     a TFE-fluorocarbon-sealed screw  cap bottle and store at  room
     temperature.  Addition of 5 fil of the  internal standard  fortifying
     solution to 5 mL of sample extract  results in  a final internal
     standard concentration of 0.1 /zg/mL.  Solution should be replaced
     when ongoing  QC (Sect. 9) indicates a problem.  Note that PCNB has
     been shown to be an effective internal standard for the method
     analytes, but other compounds may be used if the quality control
     requirements  in Section 9 are met.

7.12 SURROGATE STANDARD SOLUTION — Prepare a surrogate standard
     fortifying solution by accurately weighing approximately 0.0050 g of
     pure DCB.  Dissolve the DCB in MTBE and dilute to volume in a 10-mL
     volumetric flask.  Transfer the  surrogate standard fortifying
     solution to a TFE-fluorocarbon-sealed screw cap bottle and store at
     room temperature.  Addition of 50 /zL of the surrogate standard
     fortifying  solution to a 1-L sample prior to extraction results in
     a surrogate standard concentration  in the sample of 25 //g/L and,
     assuming quantitative recovery of DCB, a surrogate standard
     concentration in the final extract of 5.0 tig/ml.  Solution should be
     replaced when ongoing QC  (Sect. 9)  indicates a problem.  Note DCB
     has been shown to be an effective surrogate standard for the method
     analytes, but other compounds may be used if the quality control
     requirements  in Section 9 are met.

7.13 LABORATORY PERFORMANCE CHECK SOLUTION -- Prepare by accurately
     weighing 0.0010 g each of chlorothalonil, chlorpyrifos, DCPA, and
     HCH-delta.   Dissolve each analyte in MTBE and dilute to volume in
     individual  10-mL volumetric flasks.  Combine 2 /zL of the
     chloropyrifos stock solution, 50 (il of the DCPA stock solution,  50
     fil of the chlorothalonil  stock solution,  and 40 ill of the HCH-delta
     stock solution to a 100-mL volumetric flask and dilute to volume
     with MTBE.   Transfer to a TFE-fluorcarbon-sealed screw cap bottle
     and store at room temperature.   Solution should be replaced when
     ongoing QC indicates a problem.


                               508-10

-------
     7.14 GO DEGRADATION CHECK SOLUTION — Prepare a solution in MTBE
          containing .endrin and 4,4'-DDT each at a concentration of 1 fig/ml.
          This solution will be injected to check for undesirable degradation
          of these compounds in the injection port by looking for endrin
          aldehyde and endrin ketone or for 4,4'- DDE and 4,4'- ODD.
              i
8.   SAMPLE COLLECTION. PRESERVATION. AND STORAGE

     8.. 1  Grab samples must be collected in glass containers.  Conventional
          sampling practices (6) should be followed; however, the bottle must
          not be prerinsed with sample before collection.

     8.2  SAMPLE PRESERVATION

          8.2.1   If residual chlorine is present, add 80 mg of sodium
                 thiosulfate per liter of sample to the sample bottle prior to
                 collecting the sample.

          8.2.2   After adding the sample to the bottle containing sodium
                 thiosulfate, seal the sample bottle and shake until
                 dissolved.  .

          8.2.3   Samples must be iced or refrigerated at 4°C from the time of
                 collection until  extraction.  Preservation study results
                 indicate that most of the target analytes present in the
                 samples are stable for 7 days when stored under these
              . •   conditions.  Preservation data for the analytes
                 chlorthalonil, alpha-HCH, delta-HCH, gamma-HCH, cis-
                 permethrin, trans-permethrin, and trifluralin are
                 nondefinitive, and therefore if these are analytes of
                 interest, it is recommended that the samples be analyzed
                 immediately.  Analyte stability may be affected by the
                 matrix; therefore, the analyst should verify that the
              ,   preservation technique is applicable to the- samples under
              •   study.                                      '    '• *

          8.2.4   All performance data presented in this method are from
                 samples preserved with mercuric chloride.  No suitable
                 preservation agent (biocide) has been found other than
                 mercuric chloride.  However, the use of mercuric chloride is
                 not required due to its toxicity and potential  harm to the
                 environment.

          8.2.5   In 'some circumstances where biological degradation of target
                 pesticides might be expected, use of mercuric chloride may be
                 appropriate to minimize the possibility of false-negative
                 results.  If mercuric chloride is to be used,  add it to the
        •.        sample bottle in amounts to produce a concentration of
                 10 mg/L.  Add 1 mL of a solution containing 10 mg/mL of
                 mercuric chloride in reagent water to the sample bottle at
                 the sampling site or in the laboratory before shipping to the
                 sampling site.  A major disadvantage of mercuric chloride is

                                    508-11

-------
                 that it is a highly toxic chemical; mercuric chloride must be
                 handled with caution, and samples containing mercuric
                 chloride must be disposed of, properly.

     8.3  EXTRACT STORAGE

          8.3.1  Sample extracts should be stored at 4°C away from light.  A
                 14-day maximum extract storage time is recommended.  However,
                 analyte stability may be affected by the matrix; therefore,
                 the analyst should verify appropriate extract holding times
                 applicable to the samples under study.

9.   QUALITY CONTROL

     9.1  Minimum quality control (QC) requirements are initial demonstration
          of laboratory capability,  determination of surrogate compound
          recoveries in each sample and blank, monitoring internal standard
          peak area or height in each sample and blank (when internal standard
          calibration procedures are being (employed), analysis of laboratory
          reagent blanks, laboratory fortified samples, laboratory fortified  •
          blanks, and QC samples.  An MDL for each analyte must also be
          determined.

     9.2  Laboratory Reagent Blanks  —  Before processing any samples, the
          analyst must demonstrate that all  glassware and reagent
          interferences are under control.  Each time a set of samples is
          extracted or reagents are  changed,  a laboratory reagent blank (LRB)
          must be a'nalyzed.  If within the retention time window of any
          analyte of interest the LRB produces a peak that would prevent the
          determination of that analyte,  determine the source of contamination
          and eliminate the interference  before processing samples.

     9.3  INITIAL DEMONSTRATION OF CAPABILITY

          9-.3.1  Select a representative  fortified concentration (about 10
                 times EDL or at a concentration that represents a mid-point
                 of the calibration  range for each analyte.   Prepare a primary
                 dilution standard (in methanol) containing each analyte at
                 1000 times selected concentration.   With a syringe,  add 1 mL
                 of the concentrate  to each:of four to seven 1-L aliquots of
                 reagent water,  and  analyze  each of these LFBs according to
                 procedures beginning in  Section 11.

          9.3.2  For each analyte, the recovery value for all  replicates must
                 fall  in the range of R ± 30% using the value for R
                 demonstrated  for reagent water in Table  2.   The precision of
                 the measurements, calculated as relative standard deviation
                 (RSD),  must be  20%  or less.   For those compounds that fail
                 these criteria,  this procedure must  be repeated using four
                 fresh samples  until  satisfactory performance has been
                 demonstrated.
                                    508-12

-------
     9.3.3  For each analyte, determine the MDL.  Prepare a minimum of 7
            LFBs at a low concentration.  Fortification concentration in
            Table 3 may be used as a guide, or use calibration data
            obtained in Section 10 to estimate a concentration for each
            analyte that will produce a peak with a 3-5 times signal to
            noise response.  Extract and analyze each replicate according
            to Sections 11 and 12.  It is recommended that these LFBs be
            prepared and analyzed over a period of several days, so that
            day to day variations are reflected in the precision data.
            Calculate mean recovery and standard deviation for each
            analyte. Use the equation given in Table 3 to calculate the
            MDL.

     9.3.4  The initial demonstration of capability is used primarily to
            preclude a laboratory from analyzing unknown samples via a
            new, unfamiliar method prior to obtaining some experience
            with it.  It is expected that as laboratory personnel gain
            experience with this method the quality of data will improve
            beyond those required here.

9.4  The analyst is permitted to modify GC columns, GC conditions,
     concentration techniques (i.e. evaporation techniques), internal
     standards or surrogate compounds.  Each time such method
     modi.fications are made, the analyst must repeat the procedures in
     Section 9.3.

9.5  ASSESSING SURROGATE RECOVERY

     9.5.1  When surrogate recovery from a sample or method blank is <70%
            or >130%, check calculations to locate possible errors,
            fortifying solutions for degradation, contamination or other
            obvious abnormalities, and instrument performance.  If those
            steps do not reveal  the cause of the problem, reanalyze the
            extract.

     9.5.2  If a LRB extract reanalysis fails the 70-130% recovery
            criterion, the problem must be identified and corrected
            before continuing.

     9.5.3  If sample extract reanalysis meets the surrogate recovery
            criterion, report only data for the reanalyzed extract.  If.
            sample extract reanalysis continues to fail the surrogate
            recovery criterion,  report all data for that sample as
            suspect.

9.6  ASSESSING THE INTERNAL STANDARD

     9.6.1  When using the internal standard calibration procedure, the
            analyst must monitor the IS response (peak area or peak
            height) of all samples during each analysis day.  The IS
            response for any sample chromatogram should not deviate from


                               508-13

-------
     9.6.2
the daily calibration check standards IS response by more
than 30%.

If >30% deviation occurs with an individual  extract, optimize
instrument performance and inject a second aliquot of that
extract.                                      •     •
            9.6.2.1  If the reinjected aliquot produces an acceptable '
                     internal standard response report results for that
                     aliquot.

            9.6.2.2  If a deviation of greater than 30% is obtained for
                     the re-injected extract, analysis of the sample
                     should be repeated beginning with Section 11,
                     provided the sample is still available.  Otherwise,
                     report results obtained from the re-injected
                     extract, but annotate as sUspect.

     9.6.3  If consecutive samples fail the IS response acceptance
            criterion, immediately analyze a calibration check standard.

            9.6.3.1  If the check standard provides a response factor
                     (RF) within 20% of the predicted value, then follow
                     procedures itemized in Section 9.6.2 for each sample
                     failing the IS response criterion.

            9.6.3.2  If the check standard provides a response factor
                     which deviates more than 20% of the predicted value,
                     then the analyst must recalibrate, as specified in
                     Section 10.  After calibration is restored,
                     reanalyze sample extracts that failed Sect 9.6.2
                     criteria.

9.7  ASSESSING LABORATORY PERFORMANCE - LABORATORY FORTIFIED BLANK

     9.7.1  The laboratory must analyze at least one laboratory fortified
            blank (LFB) sample with every twenty samples or one per
            sample set (all s'amples extracted within a 24-h period)
            whichever is greater.  The fortified concentration of each
            analyte in the LFB should be 10 times EDL or a concentration
            that represents a mid-point of the calibration range.
            Calculate accuracy as percent recovery (X,).   If the recovery
            of any analyte falls outside the control limits (see Sect.
            9.7.2), that analyte is judged out of control, and the source
            of the problem should be identified and resolved before
            continuing analyses.

            Note: It is suggested that one multi-component analyte
            (toxaphene, chlordane or an Aroclor) LFB also be analyzed
            with each sample set.  By selecting a different multi-
            component analyte for this LFB each work shift,  LFB data can
                               508-14

-------
            be obtained for all of these analytes over the course of
            several days.

     9.7.2  Until sufficient data becomes available from within their own
            laboratory, usually a minimum of results from 20 to 30
            analyses, the laboratory should assess laboratory performance
            against the control limits in Sect. 9.3.2 that are derived
            from the data in Table 2.  When sufficient internal
            performance data becomes available, .develop control limits
            from the mean percent recovery (X) and standard deviation (S)
            of the percent recovery.  These data are used to establish
            upper and lower control limits as follows:

                     UPPER CONTROL LIMIT  - X + 3S
                     LOWER CONTROL LIMIT  = X - 3S

            After each five to ten new recovery measurements, new control
            limits should be calculated using only the most recent 20-30
            data points.  These calculated control limits should not
            exceed those established in Sect. 9.3.2.

     9.7.3  It is recommended that the laboratory periodically document
            and determine its detection limit capabilities for the
            analytes of interest.

     9.7.4  At least quarterly, analyze a QC sample from an outside
            source.

9.8  ASSESSING METHOD PERFORMANCE - LABORATORY FORTIFIED SAMPLE MATRIX

     9.8.1  The laboratory must add a known concentration to a minimum of
            10% of the routine samples or one sample per set, whichever
            is greater.  The added concentration should not be less then
            the background concentration of the sample selected for
            fortification.  Ideally, the fortified analyte concentrations
            should be the same as that used for the LFB (Section .9.7).
            Over time, samples from all routine sample sources should be
            fortified.

     9.8.2  Calculate the percent recovery, P, of the concentration for
            each analyte, after correcting the analytical result, X, from
            the fortified sample for the background concentration, b,
            measured in the unfortified sample, i.e.,:

            P = 100  (X - b) / fortifying concentration,

            and compare these values to reagent water recoveries listed
            in Table 2.  The calculated value of P must fall in the range
            of R ± 35%.  If P exceeds this control limit, and the
            laboratory performance for that analyte is shown to be in
            control  (Sect. 9.7), the recovery problem encountered with
            the dosed sample is judged to be matrix related, not system

                               508-15

-------
             related.   The  result  for  that  analyte  in  the unfortified
             sample  is  labeled  suspect/matrix to  inform the data user that
             the  results  are  suspect due to matrix  effects.

9.9  ASSESSING  INSTRUMENT  SYSTEM  - LABORATORY  PERFORMANCE CHECK (LPC)

     9.9.1   Laboratory performance check  (LPC).  After initial
             demonstration  of capability,  instrument performance should be
             monitored  on a daily  basis by  analysis of the LPC sample.
             The  LPC sample contains compounds  designed to monitor
             instrument sensitivity, column performance (primary column)
             and  chromatographic performance.   LPC  sample components and
             performance  criteria  are  listed in Table  4.  Inability to
             demonstrate  acceptable instrument  performance indicates the
             need for reevaluation of  the  instrument system.  The
             sensitivity  requirements  are  set based on the EDLs published
             in this method.  If laboratory EDLs  differ from those listed
             in this method,  concentrations of  the  LPC standard compounds
             must be adjusted to be compatible  with the laboratory EDLs.

     9.9.2   Degradation  of DDT and endrin  caused by active sites in the
             injection  port and GC columns may  occur.  This should be
             checked on a daily basis  by injecting  the GC degradation
             check solution.  Look for the degradation products of 4,4'-
             DDT  (4,4'-DDE  and 4,4'-DDD) and the  degradation products of
             endrin  (endrin aldehyde,  EA and endrin ketone, EK).  For
             4,4'-DDT,  these  products will elute  just  before, the parent,
             and  for endrin,  the products will  elute just after the
             parent.  If degradation of either  DDT  or  endrin exceeds 20%,
             resilanize the injection  port liner  and/or break off a meter
             from the front of the column.  The degradation check solution
             is required each day  in which analyses are performed.

            % degrade  = Total  DDT degradation peak area (DDE+DDD)  X100
             of 4,4'DDT     Total DDT peak area  (DDT+DDE+DDD)

            % degrade  = Total  EA + EK peak area  X100
             of endrin        Total endrin-fEA+EK  area

             NOTE: If the analyst can verify that 4,4 DDT, endrin, their
             breakdown  products, and the analytes in the IPC solution are
             all resolved,  the IPC solution and the GC degradation check
             solution may be  prepared and analyzed as a single solution.

9.10 The laboratory may adopt additional  quality control  practices for
     use with this method.   The specific practices that are most
     productive depend upon the needs of the laboratory and the nature of
     the samples.  For example, field or laboratory duplicates may be
     analyzed to assess the precision of the environmental  measurements
     or field reagent blanks may be used to asses contamination of
     samples under site conditions, transportation and storage.


                               508-16

-------
10.   CALIBRATION

     10.1 Establish GC operating parameters equivalent to those .indicated in
          Sect.  6.8.  The GC system must be calibrated using the internal
          standard technique (Sect. 10.2) or the external standard technique
          (Sect. 10.3).  Perform, the endrin and DDT degradation check
          described in Sect. 9.9.2. If degradation of either DDT or endrin
          exceeds 20%, take corrective action before proceeding with
          calibration.

     10.2 INTERNAL STANDARD CALIBRATION PROCEDURE — To use this approach, the
          analyst must select one or more internal standards compatible in
          analytical behavior to the compounds of interest.  The analyst must
          further demonstrate that the measurement of the internal  standard is
          not affected by method or matrix interferences.  PCNB has been
          identified as a suitable internal  standard.  Data presented in this
        •  method were generated using the internal standard calibration
          procedure.

          10.2.1 Prepare calibration standards at a minimum of three
                 (recommend five)  concentration levels for each analyte of
                 interest by.adding volumes  of one or more stock standards to
                 a volumetric flask.  To each calibration standard,  add a
                 known constant amount of one or more of the internal
                 standards,  and dilute to volume with MTBE.   Guidance on the
                 number of standards is as follows:   A minimum of three
                 calibration standards are required to calibrate  a  range of a
                 factor of 20 in concentration.   For a factor of 50,  use at
                 least four standards, and for a factor of 100,  at  least five
                 standards.   The lowest standard should represent analyte
                 concentrations near,  but above,  their respective EDLs.   The
                 remaining standards should  bracket the analyte concentrations
                 expected in the sample extracts,  or should define  the  working
                 range of the detector.   The calibration standards  must
                 bracket the analyte concentration found in the sample
                 extracts.   NOTE:   Calibration standard solutions must  be
                 prepared such  that no unresolved  analytes are mixed  together,
                 and calibration standards for toxaphene,  chlordane  and each
                 of the Aroclors must  be prepared  individually.

          10.2.2 Analyze each calibration standard according to the  procedure
                 (Sect.  11.4).   Tabulate response  (peak height or area)
                 against concentration for each  compound and internal
                 standard.   Calculate  the response factor (RF)  for  each
                 analyte and surrogate using Equation  1.   RF is a unitless
                 value.

                                (A8)(Cis)
                       RF =  	             Equation  1
                                (A,.)(C.)

                 where  :

                                    508-17

-------
                = Response for the analyte to be measured.
                = Response for the internal standard.
                = Concentration of the internal  standard (p.g/1).
                = Concentration of the analyte to be measured
             Note:  For  options  on  calculating  response  factors  for multi-
             component  analytes refer  to  Sect.  12.4.

      10.2.3  If  the  RF  value  over  the  working  range  is  constant  (20%  RSD
             or  less) the  average  RF can  be  used  for  calculations.
             Alternatively, the results can  be  used  to  plot  a calibration
             curve  of response  ratios  (As/Ajs)  vs.  Cs.

      10.2.4  The working calibration curve or  calibration  factor must  be
             verified on each working  day by the measurement of a minimum
             of  two  calibration check  standards, one  at the  beginning  and
             one at  the end of  the  analysis  day.   These check standards
             should  be  at  two different concentration levels to verify the
             calibration curve.  For extended  periods of analysis (greater
             than 8  hrs.), it is strongly recommended that check standards
             be  interspersed with  samples at regular  intervals during the
             course  of  the analyses.   If  the response for  any analyte
             varies  from the predicted response by more than ±20%, the
             test must  be  repeated  using  a fresh calibration standard. ...If
             the results still  do not  agree, generate a new calibration
             curve.  For those  analytes that failed the calibration
             verification, results  from field  samples analyzed since the
             last passing  calibration  should be considered suspect.
             Reanalyze  sample extracts for these analytes  after acceptable
             calibration is restored.  WARNING: A dirty injector insert
             will cause poor sensitivity  for the late eluting analytes.

             NOTE:   It  is  suggested that  a calibration verification
             standard of one multi-component analyte also  be analyzed each
             day or work shift.  By alternating the selection of the
             multi-component analyte chosen, continuing calibration data
             can be obtained for all of these analytes over a period of
             several days.

     10.2.5  Verify calibration standards periodically,  recommend at least
             quarterly,  by analyzing a standard prepared from reference
             material obtained  from an independent source  (QCS).  Results
             from these analyses must be  within the limits used to
             routinely check calibration  (Sect. 10.2.4).

10.3 EXTERNAL STANDARD CALIBRATION PROCEDURE

     10.3.1  Prepare calibration standards as in Section 10.2.1, omitting
            the use of the internal standard.                           •

     10.3.2 Starting with the  standard of lowest concentration, analyze
            each calibration standard according to Sect.  11.4 and

                               508-18

-------
                 tabulate response  (peak height or area) versus the
                 concentration in the standard.  The results can be used to
                 prepare a calibration curve for each compound.
                 Alternatively, if the ratio of response to concentration
                 (calibration factor) is a constant over the working range
                 (20% RSD or less), linearity through the origin can be
                 assumed and the average ratio or calibration factor can be
                 used in place of a calibration curve.

                 Note: For options on calulating a calibration factor for
                 multi-component analytes,  refer to Sect. 12.4.

          10.3.3 The working calibration curve or calibration factor must be "
                 verified on each working day by the procedures described in
                 Section 10.2.4.   Note:  It is suggested that one multi-
                 component analyte calibration be verified each day or work
                 shift.  , By alternating  the selection of the analyte (an
      .,          Aroclor or tpxaphene),  calibration verification data can be
                .obtained for all  these  analytes over a period of several
                 days.

        .  10.3.4 Verify calibration standards periodically (at least
                 quarterly),  by analyzing a QCS.

11.   PROCEDURE

     11.1  EXTRACTION (MANUAL METHOD)

          11.1.1 Mark the water meniscus on the side  of the  sample  bottle for
                 later determination of  sample volume (Sect.  11.1.6).   Add
                 preservative to  LRBs and LFBs.   Fortify the  sample with 50>L
                 of the  surrogate  standard  fortifying solution.   Pour the
                 entire  sample into a 2-L separatory  funnel.

          11.1.2 Adjust  the  sample  to pH 7  by adding  50 mL of phosphate
                 buffer.   Check pH:   add H2S04 or NaOH if necessary.

          11.1.3 Add 100  g  NaCl  to  the sample,  seal,  and shake  to dissolve
                 salt.

          11.1.4 Add 60 ml methylene chloride to  the  sample bottle,  seal, and
  ,               shake 30 s  to  rinse the inner walls.   Transfer  the solvent  to
                 the separatory funnel and  extract  the  sample fay vigorously
                 shaking  the  funnel  for  2 min with  periodic venting to  release
                 excess pressure.   Allow the  organic  layer to separate  from
                 the water phase for a minimum  of  10  min.  If the emulsion
                 interface between  layers is  more  than  one third the volume  of
                 the solvent  layer,  the  analyst must  employ mechanical
                 techniques  to  complete  the phase  separation.  The  optimum
                 technique depends  upon  the sample, but  may include  stirring,
                 filtration of  the  emulsion through glass wool,


                                    508-19

-------
            centrifugation, or other physical methods.  Collect the
            methylene chloride extract in a 500-mL Erlenmeyer flask.

     11.1.5 Add a second 60-mL volume of methylene chloride to the sample
            bottle and repeat the extraction procedure a second time,
            combining the extracts in the Erlenmeyer flask.  Perform a
            third extraction in the same manner.

     11.1.6 Determine the original sample volume by refilling the sample
            bottle to the mark and transferring the water to a 1000-mL
            graduated cylinder.  Record the sample Volume to the nearest
            5 mL.

11.2 AUTOMATED EXTRACTION METHOD — Data presented in this method were
     generated using the automated extraction procedure with the
     mechanical tumbler.

     11.2.1 Mark the water meniscus on the side of the sample bottle for
            later determination of sample volume (Sect. 11.2.6).  Add
            preservative to LRBs and LFBs.  Fortify the sample with 50 /zL
            of the surrogate standard fortifying solution.  If the
            mechanical separatory funnel shaker is used, pour the entire
            sample into a 2-L separatory funnel.  If the mechanical
            tumbler is used, pour the entire sample into a tumbler
            bottle.
              *
     11.2.2 Adjust the sample to pH 7 by adding 50 ml of phosphate
            buffer.  Check pH:  add H2S04  or  NaOH  if  necessary.

     11.2.3 Add 100 g NaCl to the sample, seal, and shake to dissolve
            salt.

     11.2.4 Add 300 ml methylene chloride to the sample bottle, seal,  and
            shake 30 s to rinse the inner walls.  Transfer'the solvent to
            the sample contained in the separatory funnel or tumbler
            bottle, seal, and shake for 10 s, venting periodically.
            Repeat shaking and venting until pressure release is not
            observed during venting.  Reseal and place sample container
            in appropriate mechanical mixing device (separatory funnel
            shaker or tumbler).  Shake or tumble the sample for 1 hour.
            Complete mixing of the organic and aqueous phases should be
            observed within about 2 min after starting the mixing device.

     11.2.5 Remove the sample container from the mixing device.  If the
            tumbler is used, pour contents of tumbler bottle into a 2-L
            separatory funnel.  Allow the organic layer to separate from
            the water phase for a minimum of 10 min.   If the emulsion
            interface between layers is more than one third the volume of
            the solvent layer, the analyst must employ mechanical
            techniques to complete the phase separation.  The optimum
            technique depends upon the sample, but may include stirring,
            filtration through glass wool, centrifugation, or other

                               508-20

-------
            physical methods.  Collect the methylene chloride extract in'
            a 500-mL Erlenmeyer flask.

     11.2.6 Determine the original sample volume by refilling the sample
            bottle to the mark and transferring the water to a 1000-mL
            graduated cylinder.  Record the sample volume to the nearest
            5 ml.

11.3 EXTRACT CONCENTRATION

     11.3.1 Assemble a K-D concentrator by attaching a 25-mL concentrator
            tube to a 500-mL evaporative flask.   Other concentration
            devices or techniques  may be used in place of the K-D if the
            requirements of Sect.  9.3 are met.

     11.3..2 Dry  the extract by pouring it through a solvent-rinsed  drying
            column containing about  10 cm of anhydrous sodium sulfate
            Collect the  extract in the K-D concentrator,  and rinse  the
            column with  20-30 ml methylene chloride;    Add  this  rinse to
            the  extract.   Alternatively,  add about  5 g anhydrous  sodium
            sulfate to dry the extract in the Erlenmeyer  flask;  swirl the
            flask to  dry extract and  allow to sit for  15  min.  Decant the
            methylene chloride extract into  the  K-D concentrator.   Rinse
            the  remaining  sodium sulfate  with two 25-mL portions  of
            methylene chloride and decant  the rinses into the  K-D
            concentrator.

     11.3.3  Add  1  to  2 clean  boiling  stones  to the  evaporative flask  and
            attach  a  macro  Snyder  column.  Prewet the  Snyder column by
         ',  adding  about 1 mL  methylene chloride  to the top    Place the
            K-D  apparatus  on  a  hot water  bath, 65 to 70°C, so  that  the
            concentrator tube  is partially immersed in the hot water  and
            the  entire lower  rounded  surface  of the flask is bathed with
            hot  vapor.  Adjust  the vertical position of the apparatus and
            the  water temperature as  required to  complete the
            concentration  in  15 to 20 min.  At the proper rate of
            distillation the balls of the column will actively chatter
            but  the chambers will  not flood.  When the apparent volume'of
            liquid reaches 2 mL, remove the K-D apparatus and allow it to
           drain and cool for at least 10 min.

    11.3.4 Remove the Snyder column  and rinse the flask and its  lower
           joint into the concentrator tube with 1 to  2 mL of MTBE.  Add
           5-10  mL of MTBE and a  fresh boiling  stone.   Attach a
           micro-Snyder  column to  the concentrator tube and prewet  the
           column by adding about  0.5 mL of MTBE to the top.  Place the
           micro K-D apparatus on  the water bath so that  the
           concentrator  tube is partially immersed in  the hot water.
           Adjust the vertical position of the  apparatus  and the water
           temperature as  required to complete  concentration in  5 to  10
           mm.   When the  apparent volume of liquid reaches  2 mL, remove
           the micro  K-D from the  bath and allow it to drain and cool.

                              508-21

-------
            Add 5-10 mL MTBE to the micro K-D and reconcentrate to 2 mL.
            Remove the micro K-D from the bath and allow it to drain and
            cool.  Remove the micro Snyder column, and rinse the walls of
            the concentrator tube while adjusting the volume to 5.0 ml
            with MTBE.

     11.3.5 Transfer extract to an appropriate-sized TFE-fluorocarbon-
            sealed screw-cap vial and store, refrigerated at 4°C, until
      f      analysis by GC-ECD.

11.4 GAS CHROMATOGRAPHY

     11.4.1 Sect. 6.8 summarizes the recommended operating conditions for
            the gas chromatogfaph.  Included in Table 1 are retention
            times observed using this method.  Other GG columns,
            chromatographic conditions, or detectors may be used if the
            requirements of Sect. 9.3 are met.

     11.4.2 Calibrate or verify the system calibration daily as described
            in Sect. 10.  The standards and extracts must be in MTBE.

     11.4.3 If the internal standard calibration procedure is used, add
            5 tJ,L of the internal standard fortifying solution to the
            sample extract, seal, and shake to distribute the internal
            standard.

     11.4.4 Inject 2 /iL of the sample extract.  Record the resulting peak
            size in area units..         .

     11.4.5 If the response for the peak exceeds the working range of the
            system, dilute the extract and reanalyze.  If internal
            standard calibration has been used, add an appropriate amount
            of additional internal standard to maintain the a
            concentration of 0.1 fjg/ml.

11.5 IDENTIFICATION OF ANALYTES

     11.5.1 Identify a sample component by comparison of its retention
            time to the retention time of a reference chromatogram.  If
            the retention time of an unknown compound corresponds, within
            limits, to the retention time of a standard compound, then
            identification is considered positive.

     11.5.2 The width of the retention time window used to make
            identifications should be based upon measurements of actual
            retention time variations of standards over the course of a
            day.  Three times the standard deviation of a retention time
            can be used to calculate a suggested window size for a
            compound.  However, the experience of the analyst should
            weigh heavily in. the interpretation of chromatograms.
                               508-22

-------
          11,5.3  Identification  requires expert judgment when sample
                  components are  not resolved chromatographically.  When GC
                  peaks obviously .represent more than one sample component
                  (i.e., broadened peak with shoulder(s) or valley between two
                  or more maxima), or any time doubt exists over the
                  identification  of a peak on a chromatogram, appropriate
                  alternate techniques, .to help confirm peak identification,
                  need to be employed.  For example, more positive
                  identification may be made by the use of an alternative
                  detector which operates on a chemical/physical principle
                  different from that originally used; e.g., mass spectrometry
                  (Method 525.2)  (1), or the use of a second chromatography
                  column.  A suggested alternative column is described in Sect.
                  6.8.

                  Note: Identify multi-component analytes by comparison of the
                  sample chromatogram to the corresponding calibration standard
                  chromatograms of chlordane,  toxaphene and the Aroclors.
                  Identification of multi-component analytes is made by pattern
                  recognition,  in which the experience of the analyst is an
                  important factor.

12.   CALCULATIONS

     12.1 Calculate analyte concentrations in the sample from the response for
          the analyte using one of the multi-point calibration procedures
          described in Sect.  10.   Do not use  the daily calibration
          verification standard to calculate  the amount of method analytes in
          samples.

     12.2 If the internal  standard calibration  procedure is used,  calculate
          the concentration (C)  in the sample using the calibration  curve or
          response factor  (RF)  determined in  Sect.  10.2 and,Equation  2.   RF is
          a  unitless  value.

                        •(A8)(i.)
          C  (tig/I) =  	              Equation  2
                       (A,S)(RF)(V0)

          where:

          As  = Response for the  parameter to be measured.
          Ajs = Response for the internal standard.
          Is  = Amount of internal standard added to each extract
          V0  = Volume of water extracted (L).

     12.3  If  the  external standard calibration procedure  is used,  calculate
          the amount  of material  injected from the  peak response  using the
          calibration  curve or  calibration factor determined  in Section  10.3
          The concentration (C)  in the  sample can be calculated from
          Equation 3.


                                   508-23

-------
          c  (/ig/L) =
          where:
                        (A)(Vt)
                        (V,)(V8)
                                       Equation 3
          :i
= Amount of material  injected  (ng)
- Volume of extract injected (fiL).
= Volume of total extract (ill).
= Volume of water extracted (ml). '
     12.4 To quantitate multi-component analytes, one of the following methods
          should be used.
          Option 1- Calculate an average response factor, calibration factor
          or linear regression equation for each multi-component analyte using
          the combined area of all the component peaks in each of the
          calibration standard chromatograms.
          Option 2- Calculate an average response factor, calibration factor
          or linear regression equation for each multi-component analyte using
          the combined areas of 3-6 of the most intense and reproducible peaks
          in each of the calibration standard chromatograms.

          When quantifying multi-component analytes in samples, the analyst
          should use caution to include only those peaks from the sample that
          are attributable to the multi-component analyte.  Option 1 should
          not be used i'f there are significant interference peaks within the
          chlordane, Aroclor or toxaphene pattern.

13.  PRECISION AND ACCURACY

     13.1 In a single laboratory, analyte recoveries from reagent water were
          used to determine analyte MDLs, EDLs (Table 3) and demonstrate
          method range.  Analytes were divided into two fortified groups for
          recovery studies.  Analyte recoveries and standard deviation about
          the percent recoveries at one concentration are given in Tables 2
          and 3.

     13.2 In a single laboratory, analyte recoveries from two standard
          synthetic ground waters were determined at one concentration level.
          Results were used to demonstrate applicability of the method to
          different ground water matrices.  Analyte recoveries from the two
          synthetic matrices are given in Table 2.

14.  POLLUTION PREVENTION

     14.1 This method uses significant volumes of organic solvents.  It is
          highly recommended that laboratories use solvent recovery systems to
          recover used solvent as sample extracts are being concentrated.
          Recovered solvents should be recycled or properly disposed of.
                                    508-24

-------
      14.2  For  information  about  pollution  prevention  that  may be  applicable to
           laboratory  operations,  consult  "Less  is  Better:   Laboratory Chemical
           Management  for Waste Reduction"  available from the  American Chemical
           Society  s Department of Government  Relations  and Science  Policy
           1155  16th Street  N.W.,  Washington,  D.C.  20036.

15.  WASTE MANAGEMENT

     15.1  It is the laboratory's  responsibility to comply  with  all  federal
           state, and local  regulations governing waste  management,  particu-
           larly the hazardous waste  identification rules and  land disposal
          restrictions.  The laboratory using this method  has the responsi-
          bility to protect the air, water, and land  by minimizing  and con-
          trolling all releases from fume hoods and bench  operations   Compli-
          ance is also required with any sewage discharge  permits and regula-
          tions.  For further information on waste management, consult "The
          Waste Management Manual for Laboratory Personnel," also available
          from the American Chemical Society at the address in Sect.  14.2.

16.   REFERENCES

     1.    "Methods for the  Determination of Organic Compounds in Drinking
          Water,  Supplement 3",  (1995).  USEPA, National  Exposure Research
          Laboratory,  Cincinnati,  Ohio,  45268.

     2.    ASTM  Annual  Book  of Standards,  Part  11,  Volume 11.02  D3694-82
          "Standard Practice for  Preparation of Sample Containers  and for
          Preservation,"  American  Society  for  Testing  and Materials, Philadel-
          phia,  PA,  1986.

     3.    "Carcinogens  -  Working  with Carcinogens,"  Department of  Health
          Education, and  Welfare,  Public Health  Service, Center  for  Disease
          Control,  National  Institute for Occupational Safety  and  Health
          Publication  No. 77-206,  Aug.  1977.                    ,         '

     4.    "OSHA Safety  and  Health  Standards, General Industry  "  (29  CFR 1910)
         Occupational  Safety and  Health Administration, OSHA  2206,  (Revised
         January  1976).                                             v^cviicu,

     5.   "Safety  in Academic Chemistry Laboratories," American  Chemical
         Society Publication, Committee on  Chemical Safety, 3rd Edition,


    6.   ASTM Annual Book of Standards, Part  11, Volume 11.01, D3370-82
          Standard Practice for Sampling Water," American  Society for Testing
         and Materials, Philadelphia, PA,  1986.                             y
                                   508-25

-------
17.  TABLES. DIAGRAMS. FLOWCHARTS AND VALIDATION DATA

                 TABLE 1. RETENTION TIMES FOR METHOD ANALYTES
                                            Retention Time3
                                               (minutes)
                                      Primary          	Alternative
Etridiazole
Chlorneb
Propachlor
Trifluralin
HCH-alpha
Hexachlorobenzene
HCH-beta
HCH-gamma
PCNB (internal std.)
HCH-delta
Chlorthalonil
Heptachlor
Aldrin
Chlorpyrifos
DC PA
Heptachlor epoxide
Chlordane-gamina
Endosulfan I
Chlordane-alpha
4,4'-DDE
Dieldrin
Endrin
Endosulfan II
Chi orobenzi late
4,4'-DDD
Endrin aldehyde
Endosulfan sulfate
4, 4' -DDT
Methoxychlor
cis-Permethrin
trans-Permethrin
DCB
23.46
25.50
28.90
31.62
31.62
31.96
33.32
33.66
34.00
35.02
35.36
37.74
40.12
40.6
41.14
42.16
43.52
44.20
44.54
45.90
45.90
46.92
47.60
47.94
48.28
48.62
49.98
50.32
53.38
58.48
58.82
64.10
22.78
26.18
30.94
(b)
32.98
(b)
40.12
35.36
34.00
41.48
39.78
36.72
38.08
(b)
41.14
42.16
43.86
43.52
44.54
44.88
45.90
(b)
51.68
48.28
46.92
46.92
49.30
50.32
53.72
(b)
(b)
(b)
   Columns and analytical conditions are described in Sect. 6.8.1 and 6.8.2,
   Data not available.
                                     508-26

-------
g
a
o
§ °CM
 +j
QJ
s ^
^^
i
Z -a
1 1 1 , 	 ,
CD
«* S- c
UJ CU CO
tT5
^^ "^C
O
Q£ O

< CO
•JF

O (O
Z OJ
>_ ^
ce:
o
o -o
«C O)
>• £ 0 — J
^^ | > Q Q
1— S- <_> =J
S £
o
oa
_i
UJ
HH
CVJ CU
UJ >
1— •<






: ocncnncsjcNjr-H^Hcn^'^-^i-cM'^-oja:
r-H

O*l O^ O"l LO CSI r— 1 r—H i— 1 O*i ^" C\J ^}" CM ^~ ^" *^C
r-H i— i r— ) r— (


oiouni^cocootOLOoorocNjfocoooco
r-HCNJCXJlOOOrOOOCNJr-HCTlCO^-lCri^HO
' .


^D O*i O1 O"! CTl ^D C3 O^ O*t OS CD ^D O^ CTi CD CT
r-H r-H i— H t— 1 t— I i— H



. LD ai cr> 10 ^ CNJ 
fO fO CU 4— CU
-c e +j , — -a
Q. E 03 i — 3 >I"-H
fO CT> *i — C CU
OJ OJ C i — nJfCi — ro
CCJ3CU(O QUJI — CM— <+- rci+-
•i— s- i- s- i. j_ i i i -a oo to -i— -i— vi
_S-OOOOOct'- - •< i — OOS-S-O







)cni~~ocNjr--.p-~oooocviLnr-Hioio
»roi^u3ior-.i-~cr,«;Mocy,i-.r--w

r-H r-H r— ( ,-H


•^^Lncna^r-Hcooor-Hini^voco
<£>r^ir)r-HU3r-«!t-0
r— 1 i — 1 i — 1 ,— | ,_)


(nocricDr— icoocooc3^cr>r-Hoo
|~H f— H r—H i-H r-H


cvi ' — * r — CNJ oo oo i — oo f-~ i — i r-*. co csj
l£>OlQi — lOli — lOOr-H CO CTl l^> CTl LO
' — 1 r-H r-H CM i-H


*™H r-H t-H r-H I— H r-H



oooooooooLouoLno
CU
T3 CU
X CU •<-
O M c. S~
Q- c -r- ^r
CU -Q O .C CU C
•— S-S-Or— -UES.T-
OfO rOrOOOi-^ICUS-Oi —
N-CoJ+JEi — i — ouECUi — oJ
"3 Q--MI — E-C-Ci — >>S_Q-jr S-
"OrO-Q"OCTifOn3(JOQ-i/ir8r^
••— 1 1 1 1 4-> -4-> to JZ 1 C Q.4-
i-innzinin Q-Q-x-t-" 
-------























•o
cu
3
™
•r™
4"*
C
o
C-J



CNI
III
L.I.I
f
CQ
I--I-
t^





































•
t/J
0)

Q.
£
CO
00

CO

1 —

(f_
0

c:
OJ
cu
PS

CU
4->
4->
c
QJ
00
cu
s-
CL
cu
S-
•a
C
nj
_^
c
OJ
r— •
f>

C


T3
CU
CJ
CU
4_>
cu


4->
c
3
O
03

S-
o
<4-
TD
CU
O
CU
S-
o
u

OJ

03
O































•
r
CU
o
o
QJ


.j
,—
CU
o

cu
CL
cu

>J 4-3
s- ^
CU 'i
> . o
0 ,
CJ j-
£ °
J_J
-!-> «
t= .,_
cu >
CJ o>
ol -°
°- ~s

O

s-
0)
4-9
O3
3

cn
£1
•r—
i-
CL
CO

c
03
QJ
•t-J
S-
O)
s-
3
4^
03
•z.

QJ
3
CL
O
00
f*
^^

. n
.^^
C
03
[^
•
C g~
•i- 03
cn
-o ••-
c jr
3 O
O •!-
4- S
4-^ •*
C -C
3 4->
O 3
E 0
03 E
o cl
<4-
•o -^
CU
4J ^)
CJ C
CU 03
S- CL
S- E
O O
C_) <_>
e 1 mg/L concentration
bstances Society
J= 3 T3
4J CO QJ
00
4J CJ 3
03 -i—
S CO
T3 3 03
•r- 3= 3
O
03 i — -
OJ .-— *.
CJ C O
•r- 0 T3
> -r- OJ
r— +J S-
3 OJ O
<4- E r—
S- 0
t— QJ c_5
4-^ 4-^
•r— C" »*
3 i-< S-
QJ
TD QJ >
QJ ^1 S=
•r- +J Q)
4— O
•r— E
. 4-J O £=
S- S- -r-
0 4-

O) CU
S- r— >
O) -Q i-
4-J OJ 3
OJ i — CO
s-.-
OJ r—
4-> > OJ
S= OJ CJ
QJ -i—
cn •* cn
03 T3 O
QJ ••— i —
S- O O
03 QJ
o
•^^ O
CZ -i— 00
03 > QJ
.Q 3 03
c^ +j
C CO
•i- -a
QJ -O
-o N QJ
GI «r^ ^J
3 S- -i-
O CU C
CJ
4-> 03 QJ
C S- -C
3 03 4J
O -C
E 0 -C
03 1 4->
O CU
4- 3 -d
cu
CU 03
4-> •!-
CJ • O
CU i — O
S- QJ 00
S- > 00
O QJ 03
O i — ^-^
508-28

-------
                               TABLE 3.
SINGLE LABORATORY ACCURACY, PRECISION, METHOD DETECTION LIMITS (MDLs)
AND ESTIMATED DETECTION LIMITS (EDLs) FOR ANALYTES FROM REAGENT WATER

Aldrin
Chlordane-alpha
Chlordane-gamma
Chlorneb
Chi orobenzi late
Chlorothalonil
DC PA
4,4'-DDD
4, 4 '-DDE
4,4'~DDT
Dieldrin
Endosulfan I
Endosulfan sulfate
Endrin
Endrin aldehyde
Endosulfan II
Etridiazole
HCH-alpha
HCH-beta
HCH-delta
HCH-gamma
Heptachlor
Heptachlor epoxide
Hexachl orobenzene
Methoxychlor
cis-Permethrin
trans-Permethrin
Propachlor
Trifluralin
Fortified
Cone.
//g/L
0.075
0.015
0.015
0.50
5.0
0.025
0.025
0.025
0.010
0.060
0.020
0.015
0.015
0.015
0.025
0.015
0.025
0.025
0.010
0.010
0.015
0.010
0.015
0.0050
0.050
0.50
0.50
0.50
0.025
Na
7
7
7
7
8
7
7
7
7
7
7
7
7
7
7
7
7
8
7
7
7
7
7
7
7
7
7
7
7
Recovery %
66
117
109
47
99
119
112
115
127
87
77
78
129
72
95
148
96
94
95
84
80
67
71
115
120
64
122
90
108
RSD
; %
9
8
3
34
5
12
4
5
6
23
22
25
4
18
15
35
17
8
12
7
16
7
18
43
11
24
9
18
3
MDLb
//g/L
0.014
0.0041
0.0016
0.25
2.2
0.011
0.0032
0.0044
0.0025
0.039
0.011
0.0092
0.0024
0.0062
0.011
0.024
0.013
0.0053
0.0036
0.0020
0.0060
0.0015
0.0059
0.0077
0.022
0.25
0.18
0.25
0.0026
EDLC
//g/L
0.075
0.0015
0.0015
0.5
5.0
0.025
0.025
0.025
0.01
0.06
0.02
0.015
0.015
0.015
0.025
0.024
0.025
0.025
0.01
0.01
0.015
0.01
0.015
0.0077
0.05
0.50
0.50
0.50
0.025
                               508-29

-------
a N * Number of replicates

b With these data, the method detection  limits  (MDL)  in  the  tables were
  calculated using the formula:

 MDL - S t^.-jj.gtpha . 0-99)

  whsvG*
 t, , , *lrJ,  _ ft 00, = Student's t value for the 99% confidence level with n-1
  (n"l g 1 *"3ipn3 — u»yy)

 degrees of freedom.

 n - number of replicates

 S » standard  deviation  of replicate analyses.

0 EDL = estimated detection  limit;  defined  as either MDL (Appendix B to  40 CFR
  Part 136  - Definition  and  Procedure for the Determination  of the Method
  Detection Limit -  Revision 1.11)  or a  level of compound in a sample yielding a
  peak in the  final  extract  with  signal-to-noise ratio of approximately  5,
  whichever value is higher.
                                      508-30

-------








00
+->
cu
E

S-
-^
3
CT
CU

Z
O
1— 1


CO
A (D
in
*Z r— f
OO r-4
cu -o
+-> c:
>> OJ
OJ o
C CO
OJ
CD
Cf—
O C
cu
c cu
o 2
•!-• 4«>
+-> cu
o JD
cu
|_ II +J U_
5 CU 03
Ij 00-
0 "—I
OO <-> E

^ O CD
0 . 0 =t
U4
(_>

LU
QJ
£
Q£
O
u_
eg:
uj cu
O- •*-*
Q; oj
0 £=
1— -=C
<
Q±
O
CQ


,
«*

UJ
_J
CQ
rf
5










•






O 0
CM O
o in
0 0
0 0




oo
£
^
>>
*"**
1 2
-c: o
0 Q












cu
CJ
OJ
E
£
o
s_
CU
a.
o
c—
>> §•
•I-J S-
•t— m
> 0
4-> OJ
-*-> 0
01 £= S_
CU CU .C
I— oo o
















II
-Q
O
in

O

A

C
O

•fj
3
*O
00
cu
ex:





0 0
0 0
in «d-
o o
0 0




£1
o
OJ OJ
-C -4-J
-l-> 1 —
O CU
s--o
O 1
1— :E
-C CJ
C— > ~*~































c
o
"r""
-fj
OJ
3
CT
cu
O)
-C
•i-3

a>
c:
00
3
•a
cu
•*->
OJ
1 —
3
o
OJ

(J


01
•r-

o
r-H
*^^*
i-H
2
•a
t—
oj

«^
00
O
cu
01
cz
'*~

+->
JZ
CT)
cu

1—
r"""
OJ
-C

OJ

-C
•fj
"O
2

•*
cu
a.
O)
0* +J
^^ oo
IS x->
CM
i-H
5T .
-i-3
cu ^:
t- CT)
cu ••-
-£= CU
































. .
d
o

+->
OJ
3
CT
CU
CU
_(_>

^^
-0
CU
i^~
cu
•a

00
OJ
00

03
cu
Q.
O
2
4->
cu
4->
cu
CU
cu

I «
+->
1 — II
o
00 O£
0)
a;

J3
cu
•H
«
^
^
+J
]J-J
"2
sy
OJ
cu
a.

cu
CT)
OJ
i-
cu

OJ
cu
• -C
+J
oo
•r™

2

•a
c'
OJ
00
S.X
OJ
cu
a.
0
2
+J

cu
-C
c.
cu
cu
3
+->
CU
JO
00
cu
E
•i—
+->

C
. o
•r-
+->
3

cu

c
•f" •
CO
cu ^
U OJ
c cu
cu CL
«£ 2
*4— 4->
•r*
•o cu
-c:
cu •<->
o
00
3 "~ dT
-t-> C
CU r^
s_ cu
CU 00
-C OJ
3-Q


508-31

-------
THIS PAGE LEFT BLANK INTENTIONALLY
               508-32

-------
METHOD 508.1   DETERMINATION OF CHLORINATED PESTICIDES, HERBICIDES, AND
               ORGANOHALIDES BY LIQUID-SOLID EXTRACTION AND ELECTRON CAPTURE
               GAS CHROMATOGRAPHY
                                 Revision 2.0
                  James W.  Eichelberger - Revision 1.0,  1994

                      Jean  W.  Munch - Revision 2.0,  1995
                    NATIONAL EXPOSURE RESEARCH LABORATORY
                      OFFICE OF RESEARCH  AND  DEVELOPMENT
                     U.S.  ENVIRONMENTAL PROTECTION  AGENCY
                            CINCINNATI, OHIO  45268
                                   508.1-1

-------
                                 METHOD 508.1
           DETERMINATION OF CHLORINATED PESTICIDES, HERBICIDES, AND
             ORGANOHALIDES IN WATER USING LIQUID-SOLID EXTRACTION
                    AND  ELECTRON CAPTURE GAS CHROMATOGRAPHY
1.   SCOPE AND APPLICATION

     1.1  This method utilizes disk liquid-solid extraction and gas
          chromatography with an electron capture detector to determine twenty
          nine chlorinated pesticides, three herbicides, and four
          organohalides in drinking water, ground water, and drinking water in
          any treatment stage.  Liquid solid extraction cartridges may also be
          used to carry out sample extractions.  Single laboratory accuracy,
          precision,  and method detection limit data have been determined for
          the following compounds:
          Analvte

          Alachlor
          Aldrin
          Atrazine
          Butachlor
          Chlordane-alpha
          Chlordane-gamma
          Chloroneb
          Chiorbenzilate
          Chlorthalonil
          Cyanazine
          DC PA
          4,4'-DDD
          4,4'-DDE
          4,4'-DDT
          Dieldrin
          Endosulfan  I
          Endosulfan  II
          Endosulfan  sulfate
          Endrin
          Endrin  aldehyde
          Etridiazole
          HCH-alpha
          HCH-beta
          HCH-delta
          HCH-gamma (lindane)
          Heptachlor
          Heptachlor epoxide
          Hexachlorobenzene
          Hexachlorocyclopentadi ene
         Methoxychlor
         Metolachlor
         Metribuzin
Chemical  Abstracts
Service Registry No.
      15972
        309
       1912
      23184
       5103
       5103
       2675
        510
       1897
      21725
       1861
         72
         72
         50-
         60-
        959-
      33213-
       1031-
        .72-
       7421-
       2593-
        319-
        319-
        319-
         58-
         76-
       1024-
        118-
         77-
         72-
      51218-
      21087-
-60-8
-00-2
-24-9
-66-9
-71-9
-74-2
-77-6
-15-6
-45-6
-46-2
-32-1
•54-8
•55-9
•29-3
•57-1
•98-8
•65-9
 07-8
 20-8
 93-4
 15-9
 84-6
 85-7
 86-8
 89-9
 44-8
 57-3
 74-1
 47-4
 43-5
 45-2
 64-9
                                   508.1-2

-------
                                                       Chemical  Abstracts
           Analyte                                     Service Registry No.
cis-Permethrin
trans-Permethrin
Propachlor
Simazine
Toxaphene
Trifluralin
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
54774-45-7
51877-74-8
1918-16-7
122-34-9
8001-35-2
1582-09-8
12674-11-1
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
      1.2   This  method  has  been  validated  in  a  single  laboratory  and  method
           detection  limits have been  determined  for each  analyte listed  above
           The method detection  limit  (MDL)  is  defined  as  the  statistically
           calculated minimum  amount that  can be  measured  with  99% confidence
           that  the reported value  is  greater than  zero (1).   For the  listed
           analytes (except multi-component analytes),  MDLs which range from
           0.001 /ig/L to 0.015 ng/L are  listed  in Table 3.  MDLs  for multi-
           component  analytes  (Aroclors  and toxaphene)  range from 0.01 to 0 13
2.   SUMMARY OF METHOD

     2.1  The analytes are extracted from the water sample by passing 1 L of
          sample through a preconditioned disk or cartridge containing a solid
          inorganic matrix coated with a chemically bonded C18  organic phase
          (liquid-solid extraction,. LSE).  The analytes are eluted from the
          LSE disk or cartridge with small volumes of ethyl acetate and
          methylene chloride, and this eluate is concentrated by evaporation
          of some of the solvent.  The sample components are separated,
          identified, and measured by injecting micro-liter quantities of the
          eluate into a high resolution fused silica capillary column of a gas
          chromatograph/electron capture detector (GC/ECD) system.

3.   DEFINITIONS

     3.1  INTERNAL STANDARD (IS) -- A pure analyte(s)  added to a sample,
          extract,  or standard solution in known amount(s) and used to measure
          the relative responses of other method analytes and surrogates  that
          are components of the same solution.

     3.2  SURROGATE ANALYTE (SA) - A pure analyte(s),  which is extremely
          unlikely to be found in any sample, and which 'is added to a sample
          aliquot in known  amount(s)  before extraction  or other processing,
          and is measured with the same procedures used to measure other
          sample components.   The purpose of the SA is  to monitor method
          performance with  each sample.


                                   508.1-3

-------
3.3  LABORATORY REAGENT BLANK (LRB) — A aliquot of reagent water or
     other blank matrix that is treated exactly as a sample including
     exposure to all glassware, equipment, solvents, reagents,  internal
     standards, and surrogates that are used with other samples.   The LRB
     is used to determine if method analytes or other interferences are
     present in the laboratory environment, the reagents,  or the
     apparatus.

3.4  INSTRUMENT PERFORMANCE CHECK SOLUTION (IPC) — A solution  of one or
     more method analytes, surrogates, internal standards, or other test
     substances used to evaluate the performance of the instrument system
     with respect to a defined set of method criteria.

3.5  LABORATORY FORTIFIED BLANK (LFB) — An aliquot of reagent  water or
     other blank matrix to which known quantities of the method analytes
     are added in the laboratory.  The LFB is analyzed exactly  like a
     sample, and its purpose is to determine whether the methodology is
     in control, and whether the laboratory is capable of making  accurate
     and precise measurements.

3.6  LABORATORY FORTIFIED SAMPLE MATRIX (LFM) — An aliquot of an
     environmental  sample to which known quantities of the method
     analytes are added in the laboratory.  The LFM is analyzed exactly
     like a sample, and its purpose is to determine whether the sample
     matrix contributes bias to the analytical results.  The background
     concentrations of the analytes in the sample matrix must be
     determined in  a separate aliquot and the measured values in the LFM
     corrected for  background concentrations.

3.7  STOCK STANDARD SOLUTION  (SSS) — A concentrated  solution containing
     one or more method analytes prepared  in the laboratory using assayed
     reference materials  or purchased from a reputable commercial source.

3.8  PRIMARY DILUTION STANDARD SOLUTION (PDS) — A  solution of several
     analytes prepared in the laboratory  from  stock standard solutions
     and diluted as needed to prepare calibration solutions and other
     needed analyte solutions.

3.9  QUALITY CONTROL SAMPLE (QCS)  — A solution of  method  analytes  of
     known concentrations which are used  to fortify an aliquot of LRB or
     sample matrix.  The  QCS  is obtained  from  a source external to  the
     laboratory and different  from the source  of calibration standards.
     It  is used to  check  laboratory performance with  externally prepared
     test materials.

3.10 METHOD DETECTION LIMIT (MDL)  — The  statistically calculated minimum
     amount of an  analyte that can be measured  with 99% confidence  that
     the reported  value  is greater than zero  (1).

INTERFERENCES

4.1  Method interferences may  be caused by contaminants in solvents,
     reagents, glassware, and  other  sample processing apparatus  that  lead
     to  anomalous  peaks  or elevated  baselines  in gas  chromatograms.

                               508.1-4

-------
     4.2   Interfering contamination may occur when a sample containing low
           concentrations of compounds  is analyzed immediately after a sample
           containing relatively high concentrations of compounds.  -Syringes
           and splitless injection port liners must be cleaned carefully or
           replaced as needed.  After analysis of a sample containing high
           concentrations of compounds, a laboratory reagent blank should be
           analyzed to ensure that accurate values are obtained for the next
           sample.

     4.3   It is  important that samples and standards be contained in the same
           solvent, i.e., the solvent for final working standards must be the
           same as the final solvent used in sample preparation.  If this is
           not the case, chromatographic comparability of standards to sample
           may be affected.

5.   SAFETY

     5.1   The toxicity or carcinogenicity of each chemical and reagent used in
           this method has not been precisely defined. .However, each one must
           be treated as a potential health hazard,  and exposure to these
           chemicals should be minimized.   Each laboratory is responsible for  •
          maintaining a current awareness of OSHA regulations regarding safe
           handling of the chemicals used in this method.   Additional
           references to laboratory safety are cited (2-4).,

     5.2   Some method analytes have been tentatively classified as known or
           suspected human or mammalian carcinogens.   Pure standard materials
           and stock standard solutions of these compounds should be handled
          with suitable protection to skin, eyes,  etc.

6.   EQUIPMENT AND SUPPLIES (All  specifications are suggested.   Catalog
     numbers and brand names are included for illustration only.)

     6.1  All glassware, including sample bottles,  must be meticulously
          cleaned.   This may be accomplished by washing with detergent and
          water,  rinsing with tap water,  distilled  water,  or solvent,  air-
          drying, and heating (where appropriate)  in a muffle furnace  for two
          hours  at 400°C.   Volumetric glassware must never be heated in a
          muffle furnace.

     6.2  SAMPLE CONTAINERS — 1-L or 1-quart amber glass  bottles fitted with
          Teflon-lined screw caps.   Amber bottles  are highly recommended since
          some of the method analytes are sensitive  to  light and are oxidized
          or decomposed upon exposure.

     6.3  VOLUMETRIC  FLASKS — Various  sizes.

     6.4  MICRO-SYRINGES — Various sizes.

     6.5  VIALS  --  Various  sizes  of amber vials  with  Teflon-lined screw caps.

     6.6  DRYING  COLUMN —  The drying tube  should contain  about  5 to 7 grams
          of anhydrous  sodium sulfate to  prohibit residual  water from
          contaminating  the extract.   Any small  tube  may  be  used,  such as  a


                                   508.1-5

-------
          syringe barrel, a glass dropper, etc.,  as long as no sodium sulfate
          passes through the column into the extract.

     6.7  FUSED SILICA CAPILLARY GAS CHROMATOGRAPHY COLUMN — Any capillary
          column that provides adequate resolution, capacity, accuracy,  and
          precision may be used.  A 30 m X 0.25 mm ID fused silica capillary
          column coated with a 0.25 jan bonded film of polyphen'yl methyl si li cone
          (J&W DB-5) was used to develop this method.   Any column which
          provides analyte separations equivalent to or better than this
          column may be used.

     6.8  GAS CHROMATOGRAPH — Must be capable of temperature programming,  be
          equipped for split/splitless injection, and be equipped with an
          electron capture detector.  On-column capillary injection is
          acceptable if all the quality control specifications in Sect.  9 and •
          Sect. 10 are met.  The injection system should not allow the
          analytes to contact hot stainless steel or other hot metal surfaces
          that promote decomposition.

     6.9  VACUUM MANIFOLD — A manifold system or a commercially available
          automatic or robotic sample preparation system designed for disks or
          cartridges should be utilized in this method.  Ensure that all
          quality control requirements discussed in Sect. 9 are met.  A
          standard all glass or Teflon lined filter apparatus should be used
          for disk or cartridge extraction when an automatic system is not
          utilized.

7.   REAGENTS AND STANDARDS

     7.1  HELIUM CARRIER GAS — As contaminant free as possible.

     7.2  EXTRACTION DISKS AND CARTRIDGES - Containing octadecyl bonded silica
          uniformly enmeshed in an inert matrix.  The disks used to generate
          the data in this method were 47 mm in diameter and 0.5 mm in
          thickness.  Larger disk sizes are acceptable.  The disks should not
          contain any organic compounds, either from the matrix or the bonded
          silica, that will leach into the ethyl acetate and methylene
          chloride eluant. Cartridges should be made of  inert, non-leaching
          plastic or glass, and must not leach plasticizers or other organic
          compounds into the eluting solvent.  Cartridges contain about 1 gram
          of silica or other inert inorganic support whose surface is modified
          by chemically  bonding octadecyl C18 groups.

     7.3  SOLVENTS — Methylene chloride, ethyl acetate, and methanol, high
          purity pesticide quality or equivalent.

     7.4  REAGENT WATER  — Water in which an interferant is not observed at
          the MDL of the compound of interest.  Prepare  reagent water by
          passing tap or distilled water through a filter bed containing
          activated carbon, or by using-a water purification system.  If
          necessary, store reagent water  in clean  bottles with Teflon-lined
          screw caps.

     7.5  HYDROCHLORIC ACID — 6N.
                                    508.1-6

-------
 7.6  SODIUM SULFATE — Anhydrous,  muffled at 400°C for a minimum of 4 hrs
      and stored in an air-tight clean glass container at ambient
      temperature.

 7.7  SODIUM SULFITE — Anhydrous.                              .

 7.8  PENTACHLORONITROBENZENE - >  98% purity,  for use as the internal
      standard.

 7.9  4,4-DIBROMOBIPHENYL — > 96%  purity,  for  use as the surrogate
      compound.

 7.10  STOCK STANDARD SOLUTIONS — Individual  solutions of analytes may be
      purchased;from commercial  suppliers  or prepared from pure  materials.
      These solutions are usually available at  a  concentration of 500
      Mg/mL.   These solutions are used to  make  the primary dilution
      standard.   They should be  stored in  amber vials in  a refrigerator or
      freezer.   Stock standard solutions should be replaced if ongoing
      quality control  checks indicate  a problem.

 7.11  PRIMARY DILUTION STANDARDS -  Prepare the solution(s)  to contain
      all method  analytes,  but not  the internal standard  or surrogate
      compound,  at  a concentration  of  2.5 /ig/mL in ethyl  acetate.

 7.12  INSTRUMENT  PERFORMANCE CHECK  SOLUTION --  Prepare by accurately
      weighing 0.0010  g  each of  chlorothalonil, chlorpyrifos, DCPA, and
      HCH-delta.  Dissolve  each  analyte in  MTBE and dilute to volume  in
      individual  10-mL volumetric flasks.   Combine Zjjl of the chlorpyrifos
      stock solution,  50 //L  of the DCPA stock solution,   50/yL of  the
      chlorothalonil  stock  solution, and 40//L of  the  HCH-delta stock
      solution to a  100-mL  volumetric  flask and dilute to volume  with
      ethyl  acetate.   Transfer to a  TFE-fluorocarbon-sealed  screw cap
      bottle  and  store at room temperature.  Solution  should be replaced
      when  ongoing QC  (Section 9) indicates  a problem.

7.13  CALIBRATION SOLUTIONS  - Using the primary  dilution  standards,
      prepare calibration solutions  at  six  concentrations  in ethyl
      acetate.  The  calibration  range  is dependent upon the
      instrumentation  used,  and  expected analyte  concentrations in  the
      samples to be  analyzed.  A suggested  concentration  range of
      calibration solutions  is 0.002-1.0 fjg/ml.    Note: Calibration
      standards for  toxaphene  and each  of the Aroclors must be prepared
      individually..

7.14  INTERNAL STANDARD  SOLUTION — Prepare this  solution  of
     pentachloronitrobenzene by itself in ethyl  acetate  at a
     concentration  of 10 jjg/ml.

7.15 SURROGATE COMPOUND SOLUTION — Prepare this  solution of 4,4'-
     dibromobiphenyl by itself in ethyl acetate  at a concentration of 10
    jt/g/mL.  Other surrogate compounds may be used if it can be
     demonstrated that they are not in any samples and are not interfered
     with by any analyte or other sample  component.
                               508.1-7

-------
     7.16 GC DEGRADATION CHECK SOLUTION — Prepare a solution in ethyl acetate
        \  containing endrin  and 4,4'-DDT each at a concentration of 1 ng/ml.
          This solution will be injected to check for undesirable degradation
          of these compounds in the  injection port by looking for endrin
          aldehyde and endrin ketone or for 4,4'- DDE and 4,4'- ODD.

8.   SAMPLE COLLECTION. PRESERVATION. AND STORAGE

     8.1  When sampling from a water tap, open the tap and allow the system to
          flush until the water temperature has stabilized (generally 1-2
          min).  Adjust the  flow to  about 500 mL/min and collect the sample
          from the flowing stream.   Keep sample sealed from collection time
          until analysis.  When sampling from a body of water, fill the sample
          container with water from  a representative area.  Sampling
          equipment, including automatic samplers, must not use plastic
          tubing, plastic gaskets, or any parts that may leach interfering
          analytes into the  sample.  Automatic samplers that composite samples
          over time should use refrigerated glass sample containers.

     8.2  Residual chlorine  in the sample should be reduced by adding 50 mg/L
          of sodium sulfite  (this may be added as a solid with stirring or
          shaking until dissolved, or as a prepared solution).

     8.3  Adjust the sample to pH <2 by adding 6N HC1.  It may require up txr 4
          ml to accomplish this.  It is very important that the sample be
          dechlorinated (Sect. 8.2)  before adding the acid to lower the pH of
          the sample.  Adding sodium sulfite and HC1 to the sample bottles
          prior to shipping the bottles to the sampling site is not permitted.
          HC1 should be added at the sampling site to retard any
          microbiological degradation of method analytes.

     8.4  Samples must be iced or refrigerated at 4°C from the time of
          collection until extraction.  Preservation study results show that
          the analytes (except cyanazine) are stable for 14 days in samples
          that are preserved as described in Sect. 8.2 and Sect. 8.3.
          Refrigerated sample extracts may be stored up to 30 days.

     8.5  If cyanazine is to be determined, a separate sample must be
          collected.  Cyanazine degrades in the sample when it is stored under
          acidic conditions or when  sodium sulfite is present in the stored
          sample.  Samples collected for cyanazine determination MUST NOT be
          dechlorinated or acidified when collected.  They should be iced or
          refrigerated as described  above and analyzed within 14 days.
          However, these samples must be dechlorinated and acidified
          immediately prior to fortification with the surrogate standard and
          extraction using the same  quantities of acid and sodium sulfite
          described above.

9.   QUALITY CONTROL

     9.1  Quality control requirements are the initial demonstration of
          laboratory capability followed by regular analyses of laboratory
          reagent blanks, laboratory fortified blanks, and laboratory
          fortified matrix samples.  The laboratory must maintain records to


                                    508.1-8

-------
     document the quality of the data generated.  .Additional quality
     control practices are recommended.  Determination of a MDL is also
     required.

9.2  Before any samples are analyzed or any time a new supply of disks or
     cartridges are received from a supplier, it must be demonstrated
     that a laboratory reagent blank is reasonably free of contamination
     that would prevent the determination of any analyte of concern.
     Both disks and cartridges could contain trace quantities of
     phthalate esters or silicon compounds that could prevent the
     determination of method analytes at low concentrations.  Other
     sources of background contamination are impure solvents, impure
     reagents, and contaminated glassware.  In general, background from
     method analytes should be below method detection limits.

9.3  INITIAL DEMONSTRATION OF CAPABILITY

     9.3.1  To demonstrate initial  laboratory capability,  analyze a
            minimum of four replicate laboratory fortified blanks
            containing each analyte of concern at a suggested
            concentration in the range of 0.01-0.5 ng/L.   Prepare each
            reagent water replicate by adding sodium sulfite (Sect. 8.2)
            and HC1  (Sect.  8.3) to  each sample,  then adding an
            appropriate aliquot of  the primary dilution standard
            solution(s).   Analyze each replicate according to the
            procedures described in Sect.  11.

     9.3.2  Calculate the measured  concentration of each  analyte in each
            replicate, the mean concentration of each analyte in all
            replicates,  the mean accuracy (as mean percentage of true
            value) for each analyte,  and the precision (as relative
            standard deviation, RSD)  of the measurements  for each
            analyte.

     9.3.3  For each analyte,  the mean accuracy, expressed as a
           , percentage of the true  value,  should be 70-130% and the RSD
            should be < 30%.

     9.3.4  To determine  the MDL, analyze  a minimum of seven replicate
            laboratory fortified blanks which have been fortified with
            all  analytes  of interest  at approximately 0.01 ng/L (Use  a
            higher concentration for  multi-component analytes).
            Calculate the MDL of each  analyte using the procedure
            described in  Sect.  13.2 (1).  It is recommended that these
            analyses  be  carried out over a period of three or four days
            to produce more realistic  limits.

     9.3.5  .Develop  a system of control  charts to plot the precision  and
            accuracy  of analyte and surrogate compound recoveries  as  a
            function  of time.   Charting of surrogate compound recoveries,
            which  are present  in every sample,  will  form a significant
            record of data  quality.   When  surrogate recovery from a
            sample,  a LFB,  or  a LFM is  <70% or >130%,  check calculations
            to  locate possible  errors,  the fortifying solution for
   .  .       degradation,  and  changes  in instrument  performance.   If the

                              508.1-9

-------
             cause  cannot  be  determined,  reanalyze the  sample.   If the
             surrogate  recovery  from  an  LFB  is  still  is <70% or  >130%,
             remedial action  (Sect. 10.8) will  likely be necessary.   If
             the  surrogate  recovery from a field  sample or LFM is still is
             <70% or  >130%, and  LFBs  are in  control,  a matrix effect  is
             suspected.

9.4  Assessing the Internal  Standard.   The  analyst should monitor the
     internal standard  response  (peak area  units) of all samples and LFBs
     during  each work  shift.  The IS area should not deviate from the
     latest  continuing  calibration check (Sect.  10.7) by more than 30%,
     or from the initial calibration by more than 50%.  If this criteria
     cannot  be met,  remedial  action  (Sect.  10.8) must be taken.  When
     method  performance has  been restored,  reanalyze any extracts that
     failed  Sect.  9.4  criteria.

9.5  With each group or set  of  samples  processed within a 12 hr work
     shift,  analyze  a  LRB  to  determine  background contamination.  Any
     time a  new  batch  of LSE  disks or cartridges are received,  or a new
     supply  of reagents are  used, repeat Sect. 9.2.

9.6  Assessing Laboratory  Performance.  With each group or set  of samples
     processed within  a 12 hr work shift, analyze a LFB containing each
     analyte of  interest at  a concentration of 0.01 to 0.5 /Kj/L.  If more
     than 20 samples are included in a  set, analyze a LFB for every 20
     samples.  Use the  criteria  in Sect. 9.3.3 to evaluate the  accuracy
     of the measurements.  If acceptable accuracy cannot be achieved, the
     problem must  be located  and corrected  before additional samples are
     analyzed.   Maintain control charts to  document data quality.

     Note: It is suggested that one multi-component analyte (an Aroclor
     or toxaphene) LFB  also  be analyzed with each sample set.   By
     selecting a different multi-component  analyte for this LFB each work
     shift,  LFB data can  be obtained for all of these analytes over the
     course of several   days.

9.7  Assessing Sample Matrix Effects.   In an attempt to ascertain any
     detrimental matrix effects, analyze a  LFM for each type of matrix
     (i.e. tap water, ground water,  surface water).  This need not be
     done with every group of samples unless matrices are vastly
     different.  The LFM should contain each analyte of interest at a
     concentration similar to that selected in Sect. 9.6.   Results from a
     LFM should  be within  65-135% of the fortified amount.   If these
     criteria are not met, then a matrix interference is suspected and
     must be documented.

9.8  Assessing Instrument  Performance.  Instrument performance should be
     monitored each 12  hr work shift by analysis of the IPC sample and GC
     degradation check  solution.

     9.8.1  The  IPC sample contains compounds designed to indicate
            appropriate instrument sensitivity, column performance
            (primary column)  and chromatographic performance.   IPC sample
            components   and performance criteria are listed  in Table 2.
                              508.1-10

-------
                 Inability to demonstrate acceptable instrument performance
                 indicates the need for revaluation-of-the instrument system.

          9.8.2  Inject the GC degradation check solution.  Look for the
                 degradation'products of 4,4'-DDT (4,4'-DDE and 4,4'-DDD) and
                 the degradation products of endrin  (endrin aldehyde, EA and
                 endrin ketone, EK).  For 4,4'-DDT, these products will elute
                •just before the parent, and for endrin, the products will
                 elute just after the parent.  If degradation of either DDT or
                 endrin exceeds 20%, resilanize the injection port liner
                 and/or break off a meter from the front of the column.  The
                 degradation check solution is required in each 12 hr
                 workshift in which analyses are performed.
                 % degrade
                 of 4,4'DDT

                 % degrade
                 of endrin
Total DDT degradation peak area (DDE+DDD)
Total DDT peak area (DDT+DDE+DDD)
                  X100
Total EA + EK peak area
Total endrin+EA+EK area
X100
                 NOTE: If the analyst can verify that 4,4 DDT, endrin, their
                 breakdown products, and the analytes in the IPC solution are
                 all resolved, the IPC solution and the GC degradation check
                 solution may be prepared and analyzed as a single solution.

     9.9  At least quarterly, analyze a QCS from an external source.  If
          measured analyte concentrations are not of acceptable accuracy as
          described in Sect. 9.3.3, check the entire analytical procedure to
          locate and correct the problem.

     9.10 Numerous other quality control measures are incorporated into other
          parts of this method,  and serve to alert the analyst to potential "
          problems.

10.   CALIBRATION AND STANDARDIZATION

     10.1 Demonstration and documentation of acceptable initial calibration
          are required before any samples are analyzed and is required
          intermittently throughout sample analysis as dictated by results of
          continuing calibration checks.  After initial  calibration has been
          successfully accomplished, at least one continuing calibration check
          is required each 12 hr work shift in which analyses are performed.

     10.2 Establish GC operating parameters equivalent to those below:
          Injector temperature —
          Detector temperature --
          Injection volume
          Temperature program  --
    250°C
    320°C
    2 jtiL, split!ess- for 45  sec.
    Inject at 40°C and  hold 1  min.
    program at 20°C/min.  to 160°C  hold  3  min.
    program at 3°C/min.  to  275°C with no  hold
    program at 20°C/min.  to 310°C  with  no hold
          Using the above conditions and the column in Sect.  6.7,  the total
          run time is about 50 min.   The last eluting analyte is trans-

                                   508.1-11    .

-------
     permethrin which elutes at 267°C with a retention time of 45.4 min.
     Table 1 lists all method analytes and their retention times using
     the above conditions,  ft should be noted that some method analytes
     elute very close together.  If there are unresolved peaks using the
     above temperature program, the analyst should modify the program to
     achieve resolution.

10.3 Analyze the instrument performance check sample and GC degradation
     check sample, and evaluate as described in Sect. 9.8.  If acceptance
     criteria are met, proceed with calibration.  If criteria are not
     met, take remedial action (Sect. 10.8).

10.4 Prepare calibration solutions containing all analytes of interest
     according to Sect. 7.13 in ethyl acetate.  The calibration standard
     concentrations should bracket the expected concentration range of
     each analyte in sample extracts, or define the working range of the
     detector.  Each standard must contain the internal standard,
     pentachlordnitrobenzene, at a concentration of 0.5 /zg/.mL.  The
     surrogate should also be present in each solution at that
     concentration.
     Note: Calibration standards of multi-component analytes must be
     prepared and analyzed as separate solutions

10.5 Analyze each calibration standard using the suggested
     conditions in Sect. 10'.2.  Tabulate peak area versus
     concentration for each compound and the internal standard.
     Calculate the response factor (RF) for each analyte and
     the surrogate using the following equation.
     where: As  = response for the analyte to be measured
                     AIS  = response  for  the 'internal  standard
                     CIS  = concentration of the  internal  standard  (/ig/mL)
                     C   = concentration of the  analyte to be  measured
     Note:  To calibrate for multi-component analytes, one of the
     following methods should be used.
     Option 1- Calculate an average response factor or linear regression
     equation for each multi-component analyte using the combined area of
     all the component peaks in each of the calibration standard
     chromatograms.
     Option 2- Calculate an average response factor or linear regression
     equation for each multi-component analyte using the combined areas
     of 3-6 of the most intense and reproducible peaks in each of the
     calibration standard chromatograms.

10.6 If the RF over the working range is constant (<30% RSD), the average
     RF can be used for calculations.  Alternatively, use the results to
     generate a linear regression calibration for each analyte using
     response ratios (AS/AIS) vs. Cs.
                              508.1-12

-------
     10.7 The linear regression calibration or RF must be verified on each
          work shift (not to exceed 12 hrs) by measuring .one or more
          calibration standards.  Additional periodic calibration checks are
          good laboratory practice.  It is highly recommended that an
          additional calibration check be performed at the end of any cycle of
          continuous instrument operation, so that each set of field samples
          is bracketed by calibration check standards.  Varying the
          concentration of continuing calibration standards from shift to
          shift is recommended, to evaluate the accuracy of the calibration at
          more than one point.  Calculate the RF for each analyte from the
          data measured in the continuing calibration check.  The RF for each
          analyte must be within 30% of the mean value measured in the initial
          calibration. If a linear regression calibration is being used, the
          measured amount for each analyte from the calibration verification
          test must be within 30% of the true value.  If these conditions do
          not exist, remedial action should be taken which may require
          recalibration.   For those analytes that failed the calibration
          verification, results from field samples analyzed since the last
          passing calibration should be considered suspect.  Reanalyze sample
          extracts for these analytes after acceptable calibration is
          restored.
                                                                              V
          Note:-It is suggested that a calibration verification standard of
          one multi-component analyte (an Aroclor or toxaphene)'also be
          analyzed each work shift.  By selecting a different multi-component
          analyte for this calibration verification each work shift,
          continuing calibration data can be obtained for all of these
          analytes over the course of several  days.

     10.8 The,fol1 owing are suggested remedial'actions which may improve
          method performance:

          10.8.1 Check and adjust GC operating conditions and temperature
                 programming parameters.

          10.8.2 Clean or replace the splitless injector liner.  Silanize a
                 cleaned  or new liner.

          10.8.3 Break off a short portion of the GC column from the end near
                 the injector, or replace GC column.  Breaking off a portion
                 of the column will somewhat shorten the analyte retention
                 times.

          10.8.4 Prepare  fresh calibration solutions and repeat the initial
                 calibrations.

          10.8.5 Replace  any components in the GC that permit analytes to come
                 in contact with hot metal surfaces.

11.   PROCEDURE

     11.1 DISK EXTRACTION

          11.1.1  This procedure may be carried out in the manual  mode or in
                 the automatic mode using a robotic or automatic sample

                                   508.1-13

-------
       preparation device.   If an automatic system is used to
       prepare samples, follow the manufacturer's instructions, but
       follow this procedure.  If the manual mode is used, the setup
       of the extraction apparatus described in EPA Method 525.2 (5)
       may be used.
       This procedure was developed using the standard 47 mm
       diameter disks.  Larger disks (i.e. 90 mm) may be used if
       special matrix problems are encountered.  If larger disks are
       used, the washing solvent volume is 15 ml and the elution
       solvent volumes are two 15 ml aliquots.

11.1.2 Insert the disk into  the filter apparatus or sample
       preparation unit.  Wash the disk with 5 ml of a 1:1 mixture
       of ethyl acetate (EtAC) and methylene chloride (MeCl2)  by
       adding the solvent to the disk,  then drawing it through very
       slowly to ensure adequate contact time between solvent and
       disk.  Soaking the disk may not be desirable when disks other
       than Teflon are used.

11.1.3 Add 5 mL methanol to  the disk and draw some of it through
       slowly.  A layer of methanol  must be left on the surface of
       the disk which must not be allowed to go dry from this point
       until the end of the  sample extraction.  THIS IS CRITICAL FOR
       UNIFORM FLOW AND GOOD ANALYTE RECOVERIES.

11.1.4 Rinse the disk with 5 mL reagent water by adding the water to
       the disk and drawing most through, again leaving a layer on
       the surface of the disk.

11.1.5 Add 5 mL methanol to the sample and mix well.   Mark the water
       meniscus on the side of the sample bottle for later
       determination of sample volume.

11.1.6 Add 50 jiiL of the surrogate compound solution (Sect. 7.15) and
       shake well.

11.1.7 Draw the sample through the disk while maintaining sufficient
       vacuum.  One liter of drinking water may pass through the
       disk in as little as 5 min without reducing analyte
       recoveries.  Drain the entire sample from the container
       through the disk.  Determine  the original sample volume by
       refilling the sample bottle to the mark with tap water and
       transferring the water to a 1000-mL graduated cylinder.
       Measure to the nearest 5 mL.                 •      .

11.1.8 Dry the disk by drawing air or nitrogen through the disk for
       about 10 min.

11.1.9 Remove the filtration glassware,  but do not disassemble the
       reservoir and fritted base.   Insert a collection tube into
       the vacuum manifold.   If a suction flask is being used, empty
       the water-from the flask and  insert a suitable collection
      . tube to contain the eluate.   The only constraint on the
       collection tube is that it fit around the drip tip of the
       fritted base.   Reassemble the apparatus.

                         508.1-14

-------
11.1.10 Rinse the inside walls of the sample bottle with 5 mL EtAC
        then transfer the solvent to the disk using a syringe or
        disposable pipet.  Rinse the inside walls of the glass
        filtration reservoir with this EtAC.  Draw the solvent
        through the disk very slowly to allow adequate contact time
        between disk and solvent for good analyte recoveries.

.11.1.11 Repeat the above step (Sect. 11.10) with 5 ml MeCl2.

11.1.12 Using the syringe or disposable pipet, ri.nse the filtration
        reservoir with two 3 ml portions of 1:1 EtAc:MeCl2.   Pour all
        combined eluates through the drying tube containing about 5
        to 7 grams of anhydrous sodium sulfate.  Rinse the drying
        tube and the sodium sulfate with two 3 ml portions of 1:1
        EtAc/MeCl2.   Collect all  eluate and washings in a
        concentrator tube.

11.1.13 Concentrate the extract to approximately 0.8 ml under a
        gentle stream of nitrogen while warming gently in a water
        bath or heating block.  Rinse the inside walls of the
        concentrator tube two or three times with EtAC during
        concentration.  Fortify the extract with 50 /zL of the IS
        fortifying solution (Section 7.14).  Adjust.the extract
        volume to 1.0 ml with EtAc.  •

11.1.14 Inject a 1 or 2 /uL aliquot into the gas chromatograph using
        the GC conditions used for initial calibration (Sect. 10.2).
        Table 1 lists retention times for method'analytes using these
        conditions.

11.1.15 Identify a method analyte in the sample extract by comparing
        its gas chromatographic retention time to the retention time.
        of the known analyte in a reference standard chromatogram, a
        calibration standard, or a laboratory fortified blank.  If
        the retention time of the sample peak is within the pre-
        defined retention time window, identification is considered
        positive.  The width of the retention time window used to
        make identifications should be based on measurements of
        actual retention time variations of standards during the
        course of a work shift. It is suggested that three times the
        standard deviation of the retention times obtained when the
        system is calibrated be used to calculate the window.  The
        experience of the analyst should be an important factor in
        the interpretation of a gas chromatogram.  Confirmation may
        be performed by analysis on a second column, or if
        concentrations are sufficient, by GC/MS.

        Note: Identify multi-component analytes by comparison of the
        sample chromatogram to the corresponding calibration standard
        chromatograms of toxaphene and the Aroclors.  Identification
        of multi-component analytes is made by pattern recognition,
        in which the experience of the analyst is an important
        factor.  Figures 1-8 illustrate patterns that can be expected
        from these analytes at low concentrations.  The peaks
        indicated on the chromatograms are those that were used for

                          508.1-15

-------
            quantitation.  Other peaks may be selected at the discretion
            of the analyst.

11.2 CARTRIDGE EXTRACTION

     11.2.1 This procedure may be carried out in the manual  mode or in
            the automatic mode using a robotic or automatic  sample
            preparation device.  If an automatic system is used to
            prepare samples,  follow the manufacturer's instructions,  but
            follow this procedure.   If the manual mode is used, the setup
            of the extraction apparatus described in EPA Method 525.2 (5)
            may be used.

     11.2.2 Elute each cartridge with a 5 ml aliquot of ethyl acetate
            followed by a 5 ml aliquot of methylene chloride.  Let the
            cartridge drain dry after each flush.  Then elute the
            cartridge with a 10 mL  aliquot of methanol, but  DO NOT allow
            the methanol to elute below the top of the cartridge packing.
            Add 10 mL of reagent water to the cartridge and  elute, but
            before the reagent water level drops below the top edge of
            the packing, begin adding sample to the solvent  reservoir.

     11.2.3 Pour the water sample into the 2-L separatory funnel  with the
            stopcock closed,  add 5  mL methanol  and the surrogate
            standard,  and mix well.   If a vacuum manifold is used instead
            of the separatory funnel, the sample may be transferred
            directly to the cartridge after the methanol  and surrogate
            standard are added to the sample.

     11.2.4 Drain the sample  into the cartridge being careful not to
            overflow the cartridge.   Maintain the packing material in the
            cartridge immersed in water at all  times.  After all  the
            sample has passed through the LSE cartridge,  draw air or
            nitrogen through  the cartridge for 10 min.

     11.2.5 If the setup in Method  525.2 (5) is being used,  transfer the
            125 mL solvent reservoir and LSE cartridge to the elation
            apparatus.  The same reservoir is used for both  apparatus.
            Rinse the inside  of the  separatory funnel and the sample jar
            with 5 mL ethyl acetate  and elute the cartridge  with this
            rinse into the collection tube.  Wash the inside of the
            separatory funnel and the sample jar with 5 mL methylene
            chloride and elute the  cartridge, collecting  the rinse in the
            same collection tube.  Small  amounts of residual water from
            the sample container and the LSE cartridge may form am
            immiscible layer  with the eluate. Pass the eluate through the
            drying column which is  packed with approximately 5 to 7 grams
            of anhydrous sodium sulfate and collect in a  second tube.
         •   Wash the sodium sulfate  with at least 2 mL methylene chloride
            and collect in the same  tube.  Proceed according to steps in
            Sect.  11.1.13, 11.1.14  and Sect. 11.1.15 above.
                              508.1-16

-------
 12.   DATA ANALYSIS AND CALCULATIONS

      12.1 Calculate the concentration (C)  of the analyte in the sample using
           the  response factor (RF)  determined in Sect.  10.5 and the equation
           below.

                         (As)  (Is)
           C  (/ig/L)  =	
                        (AIS)(RF)(V0)

           where:  As    = peak  area for the  analyte to be  measured
                  AjS  = peak area for the internal standard
                  Is    = amount of internal standard added  (/ig)
                  V0    = volume of water extracted (L)

           If a  linear  regression  calibration  is  used, use  the  regression
           equation  to  calculate  the amount of analyte in the sample.   All
           samples  containing  analytes.outside the calibration  range must  be
           diluted  and  reanalyzed.   When diluting, add additional  internal
           standard  to  maintain its  concentration  at 0.5  jjg/ml  in  the diluted
           extract.

      12.2  To quantitate multi-component analytes, one of'the following methods
           should  be  used.
           Option  1-  Calculate an  average response factor or linear  regression
           equation  for  each multi-component analyte using  the  combined area  of
           all the component peaks in  each  of  the calibration standard
           chromatograms.
           Option 2-  Calculate an  average response factor or linear  regression
           equation  for  each multi-component analyte using  the  combined areas
           of 3-6 of  the most  intense  and reproducible peaks i.n  each of the
           calibration  standard chromatograms.

           When quantifying multi-component analytes in samples, the analyst
           should use caution to include only  those peaks from  the sample that
           are attributable to the multi-component analyte.   Option  1 should
           not be used if there are  significant interference peaks within the
          Aroclor or toxaphene pattern.

13.   METHOD PERFORMANCE

     13.1 Method performance data was obtained using the GC column and
          conditions described in Sections 6.7 and 10.2.  Retention times are
          listed in Table 1.  All data presented here were  obtained with the
          liquid-solid extraction disk option.  Previous method development
          research has shown no significant performance  differences between
          cartridges and disks.   Method  525.2 shows  comparative recovery data
          for Method 508.1  analytes  using both cartridges and disks.

     13.2 Method detection  limits (MDL)  for all method  analytes (except
          Aroclors and toxaphene) were determined by analyzing  seven reagent
          water samples which  were fortified  with the  analytes  at a
          concentration of  0.01 /ig/L.  The  mean and  standard deviation  were
          calculated for each  analyte.  The MDL was  calculated  by multiplying
          the  standard deviation  by  the  students-t value for n-l-and a  99%

                                   508.1-17

-------
          confidence interval  (1).  The students-t value for seven  replicates
          (n-l=6)  is 3.143.   The mean recoveries and the standard  deviations
          along with the MDLs  are listed in Table 3.  Aroclor and  toxaphene
          data in  Table 3 were calculated using Option 2 in Section 12.2.

     13.3 Method accuracy and  precision were determined by analyzing two sets
          of eight reagent water samples fortified with method analytes .
          (except  the multi-component analytes) at approximately 5 and 10
          times the average MDL. The fortification concentrations  for these
          samples  were calculated by averaging the analyte MDLs and
          multiplying that average by 5 and 10.  Thus the concentrations used
          were 0.03 /ig/L and 0.048 jug/L.  Results of these analyses are listed
          in Tables 4 and 5. An additional  set of samples was analyzed at
          approximately 20 times the average MDL (0.096/zg/L).  This set of
          samples  was extracted from an artificial matrix containing Img/L
          fulvic acid.  The fulvic acid served to mimic the naturally
          occurring organic material found in many water sources.   The results
          of these analyses are listed in Table 6.

     13.4 Atrazine, hexachlorocyclopentadiene, and metribuzin appear to be
          problem analytes.   Atrazine displays low peak response when compared
          to most  of the other method analytes, and requires manual peak area
          integration even at the 0.048-/ig/L level.
          Hexachlorocyclopentadiene, while displaying relatively high peak
          response, showed poor recovery.  The resulting mean recoveries were
          50.8, 52.6, and 21.7 percent respectively for the three  levels. It
          is suspected that the higher volatility of hexachlorocyclopentadiene
          causes the problem.   Very careful, very slow nitrogen blowdown may
          produce  higher recoveries of this compound (5).
          It is suspected that Metribuzin was recovered poorly  due to
          breakthrough on C-18 media (5).

14.   POLLUTION PREVENTION

     14.1 This method utilizes liquid-solid extraction  (LSE) technology to
          remove the analytes from water.  It requires the use of very small
          volumes of organic solvent and very small quantities of pure
          analytes.  This eliminates the potential  hazards to both the analyst
          and the environment that are present when large volumes  of solvents
          are used in conventional liquid-liquid 'extractions.

     14.2 For information about pollution prevention that may be applicable to
          laboratory operations, consult "Less  Is Better:  Laboratory Chemical
          Management for Waste Reduction" available from the American Chemical
          Society's Department of Governmental  Relations and Science Policy,
          1155 16th Street N.W., Washington, D.C.,  20036.

15.   WASTE MANAGEMENT

     15.1 It is the laboratory's responsibility to  comply with all federal,
          state, and local regulations governing waste management,
          particularly the hazardous waste identification rules and land
          disposal restrictions.  The laboratory  using  this method has the
          responsibility to protect the  air, water, and land by minimizing and
          controlling all releases from  fume hoods  and  bench operations.

                                   508.1-18

-------
          Compliance is .also .required with any sewage discharge permits and
      .    regulations.  For further information on waste management, consult
  .•:.;.     "The Waste, Management Manual for Laboratory Personnel,"  also
          available from the .American Chemical Society at the address in Sect,
          14.2.

16.   REFERENCES

     1.   , J.A. Glaser, D..L. Foerst, G.D. McKee, S.A. Quave, and W.L. Budde,
          "Trace Analyses for Wastewaters," Environ. Sci. Technol.  1981 15,
          1426-1435. or 40 CFR, Part 136, Appendix B.

     2.    "Carcinogens - Working With Carcinogens," Department of  Health,
          Education,, and Welfare, Public Health Service, Center for Disease
          Control., National Institute for Occupational Safety and  Health,
          Publication No. 77-206, August 1977.

  :   3.    "OSHA Safety and Health Standards, General Industry," (29CFR1910),
          Occupational Safety and Health Administration, OSHA 2206  (Revised
          January 1976).

 .-.   4-.    "Safety in Academic Chemistry Laboratories," American Chemical
          Society Publication, Committee on Chemical Safety, 3rd Edition,
          1979. •.   •:

     5.    Munch, J. W., "Method 525.2-Determinatibn of Organic Compounds in
.-  .:.,     Drinking Water by Liquid-Solid Extraction and Capillary  Column
          Chromatography/ Mass Spectrometry" in Methods for the Determination
          of Organic Compounds in Drinking Water; Supplement 3  (1995).
          USEPA,•National Exposure Research Laboratory, Cincinnati, Ohio
          45268.
                                   508.1-19

-------
17.0 TABLES, DIAGRAMS, FLOWCHARTS, AND VALIDATION DATA
          TABLE 1.  RETENTION TIMES FOR METHOD ANALYTES USING THE GC
                    COLUMN IN SECT. 6.7 AND THE GC CONDITIONS IN
                    SECT. 10.2
ANALYTE
Hexachlorocyclopentadiene
Etridiazole
Chloroneb
Propachlor
Trifluralin
HCH-alpha
Hexachl orobenzene
Simazine
Atrazine
HCH-Beta
HCH-Gamma
HCH-Delta
Chlorthalonil
Metribuzin
Heptachlor
Alachlor
Aldn'n
Metolachlor
Cyanazine
DC PA
Heptachlor Epoxide
Chlordane-Gamma
Endosulfan I :
Chlordane-alpha
Dieldrin
4,4'-DDE
RETENTION TIME (Min)
9.64
11.41
12.39
14.69
16.29
17.01
17.44
17.86
18.23
18.33
18.71
19.21
20.27
21.88
22.78
22.86
24.81
25.02
25.21
26.49
27.20
. 28.65
29.36
29.58
30.95
31.97
                                   508.1-20

-------
TABLE 1.  RETENTION TIMES FOR METHOD ANALYTES USING THE GC
          COLUMN IN SECT. 6.7 AND THE GC CONDITIONS IN
          SECT. 10.2
ANALYTE
Endrin
Butachlor3
Endosulfan II
Chlorbenzilate
4,4'-DDD
Endrin Aldehyde
Endo.sulfan suTfate
4,4'-DDT
Methoxychlor
cis-Permethrin
trans-Permethrin
Toxaphene3
Aroclor 1016a
Aroclor 1221a
Aroclor 1232a
Aroclor 1242a
Aroclor 1248a
Aroclor 1254a
Aroclor 1260a
Pentachlorointrobenzene (IS):
4,4-Dibromobiphenyl (SUR):
RETENTION TIME (Min)
32.24
32.65
32.81
32.98
33.49
33.96
35.43
35.80
39.38
44.98
45.42
33.53, 36.48, 39.12b
18.93, 22.55, 24.83b
13.67, 18.02, 19.93b
18.93, 22.55, 24.83b
18.93, 22.55, 24.83b
24.15, 24.83, 31.40b
31.80,, 34.12, 38.88b
35.65, 41.38, 43.08b
19.02 minutes
25.64 minutes
  Retention time was determined with the following GC conditions:

  Injector temperature -- 250°C
  Detector temperature — 320°C
  Injection volume     — 2 #L, splitless for 45 sec.
  Temperature program  — Inject at 60°C and hold 1 min.
                       — program at 20°C/min.  to 160°C hold 3 min.
                       — program at 3°C/min. to 275°C with no hold
                       -- program at 20°C/min.  to 310°C with no hold
                        508.1-21

-------
TABLE 1.  RETENTION TIMES FOR METHOD ANALYTES USING THE GC
          COLUMN IN SECT. 6.7 AND THE GC CONDITIONS IN
          SECT. 10.2
   The IS retention time using these conditions is 21.15 min.
   SUR retention time using these conditions is 28.18 min.
The
   The retention times listed do not reflect the total number of
   peaks characteristic of the multi-component analyte.  Listed peaks
   indicate those chosen for quantita'tion.  Quantitative data is in
   Table 3.
                          508.1-22

-------









01
£=
CU
cu
s_
•r—
_ CT
2 S
i—
_i
o
CO

^^
f \ * — '
g £^
J3 0 en
LU
O
«=c
s:
a:
o
u.
Of
LU
o.
^ cu
O "t"
a: "*
1 5
_J


•


ro
A

^™
oo

cu"
ofanalyt
c
o
o
cu
cu
a



o
CNJ
O
0
o






01
o
14-
•r-
^1
a.
o



CM
[I
LU
_J
CO II
2
II










•4->
01
CU
1—







>>
•r~
>
-M
01
C
cu
CO


ro
LO
^™^
*
61 '

£=
re
o
CO
o
c
CU
cu
cu
JD
u_
O_



o
o
LT>
o
o








o
o











cu
e
i_
o
a.
->
O C*
s- -a
O 1
o :E














cu
ro
E
S-
c^
S-
cu
a.
i
3

O
O


















1
-t->
rO
3
CT
CU

CU
en
c:
01
-a
CU
• +j
- ro
r~*
Z7
O
r—
re
o


sJ
o
-IJ
u
ro
CVI
c: ^
ro F— i
01 3
01
3 X
ro
O CO
CO
re r— i
cu
a.

i ii
u. u_
C^ C3
Q- Q_



a
JC
4->
C
CU
-(->
4->
01
U
cu
01

c
JC
4->
T3
rO
CU
a.
cu
4J
CO

,*-**
o
f— 1
1-^-1
s'
-o
ro

•
01
O
cu
01
C
4-J
J=
CUT)
QJ
jc
<4-
re
JC

4-1
re

JC
•+—^
"O
's

re
cu
a.
cu
+->
O 01
1—* T-
3 -~-~
1— C
3 .
4_>
CU JC
S- en
cu -i-
jc cu
S JC





















^~
o
-M
ro
3
CT
CU .

cu
JC
4->

0)
<£
cu

01
re

01

ro
cu
CL
o
3
+->
cu
4^
c
cu
a>
4->
cu
JO

o
jj
•T
I— II
O
01 o;
ai
a:



J3
+J
ro
JC
4->
•a
•r™
s

sy
re
cu
a.
cu
en
re
cu
rO
CU
01
3
•a
ro

01
ro
CU
a.
O

CU
JC
4->

—
CU
cu
OJ

cu
E -
4->
C
o

4_)
^3

'aJ

c:
-r"
cu
o
cu
s_
cu
•o
cu
JC
4->
01
• pIM
3
i ^

cu
S-
cu cu
"s5









































01

re
cu
CL
o
•M
cu
<4_
o

O)
s^
cu
01
re




508.1-23

-------
LU


5
UJ
cn
V)
LU
o
D-
LU
o:
LU
s
oi
o
LU
o:
 o
 r—I
 CO



 Q




  •
 CO

 LU
 _J
 CD
a
-.
_J
§





cn
o
o
o
V

to
r-.
CO

o
CO
O
0
0

o
oo
00
o
o
CD

o
CD









o
t—
o

1"

cn
o
0
o


un
r-.
CO

CD
CO
o
o
o

CD
CO
oo
c
CD

CD
O













1""

CO
o
o
o


1 — 1


o
1 — 1
o
CD
O

O
O
CD

O
CD









CO
IM
O3



O
0


in
in

CO
CM
o
o

ID
oo
CO
0

in
CD









S-
o
o
rt
'
-J

CM
O
O
O


1 —
CO

o
CD
0
CD

CD
00
00
O
O
CD

O
O





CO


>r_
N
• C
C
^_
o

*

0
o
0


I-H
1 — 1

CM
i — I
O
CD
CD

O
r-.
o
o
o

CD
CD




CV
-C

'fO
1
cu
c
i-
o

"^~

1 — 1
0
CD
0


o
o
LTJ

CO
o
o
CD

CD
ID
O
CD
CD

CD
CD




n}
£

cr

CO
c:
a
i-
O

-*-

0
o
o


00
i-H

cn
i — i
CD
O
CD

O
CM
O
O

CD
O









J3
CO
£=
O

O

™

CM
O
0
o


o
o
i-H

o
o
o
o

o
CD
o
o

0
CD





f—

o

o

o

"

r-.
CD
0


0
CM
CM

CM
CM
CD
CD
CD

C
O
o
0

CD
CD









C
NJ

C
•; ^


O
CD
O


o
00
CM

OO
CM
O
0
o

CD
O
CD
CD
CD

O
O











Q-

— .

CO
0
o
CD


cn
oo
00

oo
o
o
o
o

o
cn
cn
o
o
o

o
CD









Q
Q
Q

^j.

.j

CO
0
0
o


G
CM
i-H
i — 1

cn
0
0
o
o

o
co
00
o
o
o

CD
o









LU
c.

^J.

^t,

o
o
CD


1 — 1
cn

r-H
CD
O
CD

CD
CM
CM
C
O

0
CD









1—
C.

^J.

^^





i-H


1 	
CM
O
O

CD
CO
co
0
0

CD
CD


^
CO
Q.
JO
o

O CO
JD rtJ
•i— CT
Q 0
1 $-

„(/•


CO
o


-------
 LU
 CS
 
=1
LU
,V ~&>
+J
^^
• "" O
O





LU
t—
_l
«c

o

CO


CO
in

ro
CM
o
o


CO
in
in
CO
0

o
CO









c
S-

c

o


o
in

CM
CO
0


00
00
0
CO
CO

CO
CO






CD
O
>•)
CO
tz
S-
T3
C

^f.

o


en
CO

in

0


CO
CO
o

o
o









a>
r—
0
IM
re

O

CO


oo
CM

ro
O
0
O


CO
I —
o
CO
0

o
CO









re
a.
re
i
0

CD
0
o
CO


CM
CM
CO

cn
CM
CO
CO
o


CO
a-i
a>
CO
CO
o

0
o









re
ai
i

0
CO
CO


in
*""*

S
CO
CO
0


o
oo
00
o
o

CO
CO









re
r—
a>
-a
i

0
o
o


to
00
CM

0
CM
CO
0
o


o
0
0
CO

CO
CO









re
E
re
en
i
CJ
_i_
in
0
O
o


co
1 — •

to
o
o
0


o
CT>
CO
o
CO

o
o









eptachlor
_i_
0
o
o


CO
in

ro
0
o
o
o


CO
^o
to
o
o
o

CO
CO



cu
~T^J
•r—
X
o
0.
LU
s_
o
r—
-C
O
re
4->
Q.
CD
^
0
O
o


CO


o
CO
0
o


CO
0
o
r— t
O
CO

CO
o




O)
sz
O)
M
sz
exachlorobe

0
o
o


in
CM
ro

ro
i-H
CO
o
0


CO
CO
CO
CO

o
o


re


CT>
o
o
o

o
CO









ISl
ZJ
i-
-I-J
cu
2:
o
o
o



-------
LU
an
*£.
O
D.
LU
LU
£
CO
CO
LU

co

LU
_J
CD
_J
o
«c
CJ
a
on

tn
S3.
.
3
a
a
fc
a
a:
en
LU
^-x
0 "
o
o
LU
«c
z

CO
1— 1
o


r— 1
if)

i-H
O
o


i-H
CO
I-H
CO
o
0
I-H



c
a>
ro
X
o
CVJ
o
o


l»-
in


o
o


CVJ
CO
IO
r— 1
O
CVJ
0


IO
I-H
o
o
o
o
S-
co
o
o


CO
10

CVI
o
o


in
CO
I-H


0


CVJ
o
u
o

-------
 LU
CD
«*
  s
eo

C


~~~.
CT
54
LU
a
a
OO
o
LU

0)
54.
-
•*






^_^
ID
F— 1



in
o
o
o



ro
o
CO
o
o











s-
0
o
ra

CM
O
CM



in
CO
o
0


01
r— 1
OO
in
CM
o
o











c:

ro
Chlorbenzi

oc
00
•—



in
o
o
o


^.
CO
oo
CM
O








re
CL
ro
Chlordane-

00
00
CM



O
o
o


CO
CM
00
in
CM
O
O








ro
E
ro
CD
I
o>
c:
ro
O
-C

CO
^t"
CM



ID
O
O
O


CO
00
OO
ID
CM
O
O











Chloroneb

in
cn
- i— <



ID
O
0
O



ID
O
I — 1
CM
CO
O
0









>r_
o
r—
ro
O
S_
O

cc
CM
I—



•=1-
o
o
0


CM
cn
cn
CM
0
o











Cyanazine

i—
•^
i —



0
o
o


CM
in
oo
CM
O
O











Q

«*
1C
i —



O
o
o


I— H
cn
oo
CM
CD
0











Q
Q
Q
1

O
O
• CM



in
o
o
0


' O
00
CO
CM
CD
O











LU
Q
Q

•e*
r-»
i — i



in
o
o


00
CO
cn
00
CM
O
o











H-
Q
Q
1
•*


1C



CM
CO
O
O


CM
CO
cn
ID
ID
0
o>
rO
Cn
O

oo
">>
O)
Q.
^
o
o
i-
-Q
Q
1

00
Lf)



C3
O
0*


in
1 — 1
cn
CM
0
o











Dieldrin

00
r-.



m
0
0


ID
CM
cn
CO
CM
O
O










1— <
Endosulfan

(C
00



m
o
0
o


cn
co
10
CM
O
0










1— I
Endosulfan
-
ir
i—



0
o



ID
O
i— (
CM
CO
0
0






0)
4-3
as
q-
•
OO
Endosulfan

o>
CO



o
o
0


in
ID
cn
CM
o
o











c
•p—
s_
c
LU
                                        508.1-27

-------
LU
ex:
   en
a
a.
LU
  CC
   LU
LU
   O


   UI
LU;
o
1/3


III
a
a
o
LU
Og

cn
=3.
2E
LU
LU
>•
_l
Z
"*









r--
CO

o
o
o

CO

CO
CT>

CD
CO
0
o






cu
-o

e-
CU
s
e
• r—
s_
•a
LU
i — i
r—

oo
0
o
o

CO

in
CT>

co
CM
O
0










cu
o
N
rO
•a
LU

«*

I1-.
o
O
O



oo
CO

10
CM
o
o











rO
O
1
1C
1C
cn
CO

o
o

o

in
CT>

00
CM
O
O











rO
CU
.a
1
re
i — i
i — i

o
o
o



1— 1
r—t
•*
CO
O
CD











-2
cu
•a
i
re
CO
1 — 1

CD
0
O



O
i — i
r—t
CO
CO
o
0











ro
£
cn
l
re
o
re

i~~

IT)
O
O
CD

LO

CD
rj>

f~-
CM
CD
CD




•o

X
o
a.
LU

o
u
a
re
t— t
Cvl

to
o
o
o

UD

in
co

CM
o
o










s_
o
u
ro
a.
cu
re

o

i
o



0
j
CM
CO
o
o




cu
c
cu
N
s=
cu
o
i-
0
_c
(J
ra
X
cu
re
oo
•*

o
o

co

CD
in

m
r— 1
O
0

ai
cu
•o
S
O-
o
r—
(J

O
o
s-
o
o
rO
X
d)
re

CD

CO
0
O
CD



CM
cn

CM
o
o









C-
o
u
X
0
cu
CM
oo

to
o
0
o



CO
1 — 1

CO
o
o










s-
o
JZ
o
o
o
'QJ
00
in

CM
o
CD
CD

in

CO

CM
1— 1
o
o










c
3
_a
• rr-
S-
o;
s:
oo
1 —

o
CD
CD

CM

CO

CO
CO
o
0







c
• r"
s_
.c
cu
s-
cu
D-
1
u

CO

o
o
o



1 — 1

CO
CO
o
0





c
•r—
&_
f~
4^
cu
cu
D_
1
01
ro
i-

r--.

o
CD
o

o

CO

oo
CM
o
0










S-
o
a
re
a.
o
a.
in
^H

CM
0
CD
CD

„

oo
to

o
CM
O
0











cu
c
IS
ro
OO
o
co

o
0
0

in

CD
00

CM
CD
O










£=
•r"
ro
s-
EJ:
•l™~
s-
                                          508.1-28

-------
LU
LU
(£.
   oo
   •cr
   o
   o
   to
   o
O
i—i
_1
a.
oa
«c
a
CO
cc
x
c
U.
a
to
o
1 1


C71
=3.
UJ
s:





UJ
>•
«c













o
ro
CM
O
0
O

J
a-i

CD
CD

















s-
o
.£!
O
re
1 —
LO


CD
0
0

00
cn
10

ro
CD
O



















c
r—
.<£
O
tO

0
CD
0

cn
£

tO
ro
o
o

















0)
£Z
N
ro
S
O
co

o
o


CD

LO
o
CD












}"
o
CM


o
0
o


to
r-H

10
LO
CD
o

















1—
a
Q
1
*"
R
UD

ro
CD
0

^
cn

*
o
a>
ro
01
o
&.
S-
oo
^
c:
CU

Q.

(~i
o

o

>rw
Q
1
*"
O
o


o
o
o


i — r
O

oo
o
o

















c
.^
• TD
cu
Q
0
LO


O
0
o


CM
CD

cn
CD
CD














, !

f-
re
t
=3
o
c
UJ
o
co
1 —

o
o
o


IO
o

1 — 1
LO
0
CD













)__(
1— t

f-
re

t
3
0
c
UJ
0
ro
cn

c
o
o


^

IO
LO
o
CD








cu
•*->
re


•^
LO

f—
re
.
3
V)
o
c
LU
0
CM
CO*

o
0
0


, —

CO
LO
o
CD



















.=
UJ
                                                 508.1-29

-------
a:
LU
a:
  oo
  •3'
  o
  : ui
s:
1-1 i
UJ
  1
  '
o
CO
o:
cn
LU
a
o
CO
o
UJ
C£
_,
=J.
z
Ul
LU


>-
_
Z














r-
o
LO
CD
o
0


LO
oo
cn

o
CD










CD

^^
^2
a.
"t;


c
t
c
LU
IO
IO
CO
O
CD
O



1 —
CD
p— <

i — 1
LO
O
0














• cu

'o
tx

•i-
fr.
S-
u.
o
o
1 — 1
LO
CD
CD
0



en
o
i — i

CXJ
LO
O
o
















ra
_c
rl
re

3:
3:
o
ID
CO
cvj
O
CD
CD


t —
O

O
CD
















re

CO
r~i

1C
0
oo
00
LO
o



0
CM
t — 1

oo
LO
o
CD
















re
-__
a
i
3:
CO
CO
1 — 1
o
CD
O



i — l
. — i
i — 1

CO
o
CD
















ro
E
£
ro
cn
l
3:
0
00
cn
o
o
o


ID
CVJ
00

o
o















J^.
o
f—
c_
re

a.
CO
3:
o
ID
ID
CO
O
0
CD



O
O
i — i

00
CD
CD







CD
T3

X
O
a.
LU

s-
o
' ' t"~
o
re

a.
CD
00
< — i
i — i
LO
O
o
0



CD

CD
CD








CD

O!
fs
c
cu
JD
O
s_
0
f__
-C
o
ro
X
O)
3:

CTl
i — l
LO
CD
CD
O


10
CvJ
Lp

LO
CvJ
CD
CD

a>
£Z
CL

•a
re
•f—
CD
a.
p

(_
^^
o
o
o
-
-C
o
ro
X
cu
3:
o
LO
cn
LO
o
0
o



cn
CD

CM
LO
o
CD













S-
o
r—
-£=
fc
X
o

a
o
IO
CO
CvJ
O
O
O



ID
CvJ

O
O
O














t»
O
-C
u
p
c
a
rrH
od
t — i
ID
O
CD
O


1 —
ID

CO
CO
O
O















C.
^,
~
^
s-
a
cn
oo
cn
CD
0
o



CXI
o

cn
CD
CD











c:
•r—
^
.C
4-^
CD
^.
CD
Q_
1
1/1
O
^
i — i
O
o
o



t — t
r- H

ID
LO
CD
CD









C
•r—
S-
^"
-t-J
CD
E
CD
' Q-
1
c/
C
re
S-
CO
CO
CO
CvJ
CD
O
CD



CVJ
O
1 — 1

cn
o
CD















5*-
p
f~
O
ra
a.
p
S-
CL.
cn
10
UD
CO
o
o
CD


CVJ
CVJ
oo

cn
CO
0
0
















Q)
c
IX
ra
E
CO
I— 1
f-

CVJ
CD
CD
0


cn
cn
oo
CO
«*•
CD
CD
O














£=
-p—
ra
^.
, 	
*-r—
1—
                                            508.1-30

-------
 o
 
Z  =4
t—i
•
«c z
o 
o
a

o
o
0
to
00
in
CD
o













Q-
O
Q
CO
c
c
CD
O
0
in
en
en
CD
0













Q
Q
Q
' 1
«*•
0
O
in
o
o
0
in
tc
in
en
CD
o













LU
Q
Q
1
*
Lf

OO
O
CD
O
0
in
00
CM
CO
o.
CD













Q
Q
1
•*"
to
to
O
CO
o
0
CM
en
to
CD
CD
O)
ro
CD
o
S-
S_
3
oo
^>
CO
Q.
"C
0
0
s_
Q
1
*
o
en
en
en
o
CD
o
CO
CM
en
en
oo
o
CD













c
il
-a
'cu
Q
o
to
oc
oo
CD
o
0
00
en
00
cc
CD












i— i
c
ro
4-
r—
3
t/1
o
T3
£1
LU
O
1 —
IO
o
o
. o
CM
en
CO
co
CD
CD












>-H
C.
CO
4-
3
VI
O
T3
C
LU
O
CO
r*"
r-.
O
O
CO
00
en
en
o
CD








cu

ro
4-
^J
OO
c
ro
<+-
3
01
O
X)
C
LU
co
C
o
o
CM
O
00
en
o
CD













s=
il
LU
                                                  508.1-31

-------
CD
LU

CIS
Dig
HI VJ
  • oo
   LU
j= O
M Z

t^g
U4 °
%
 C3
 H-<
 LU


  •
 IO

 LU

 cn
Q
CO
B£
cn
=4
*
O
a
U

LU
cn
g
LU


H-
2
•a:










o
CM
in

CO
CD
CD
O


CO
cn
r--
10
CD
CD







CD

>•
e

c
S-
-^
c.
LU
O
CO
cn

cn
CD
CD
CD


cn
CO
cn
LT>
cn
o
o










CD
O
N
"O

5_
LL
O
t--.
ID

ID
O
O,
O


•=1-
m
cn
CM
cn
CD
o











rG
.c:
rO

^
C_3
CD
CO
~

O
CD
O


CM
i — l
cn
co
cc
CD











re
<

"1
^
CD
cn
in

ID
o
0
CD


ID
cn
cn
CD
CD











re
c

«
o
CD
1 — 1
1 —

1 —
0
CD
CD



to
o
CM
CD
CD











e
fO
CJ


o
o
co
cn

o
0
o



^
co
ID
CD
0










s_
o
t
4-3
C
o>
o
co
1 —

1 —
CD
CD
O


ID
i — l
cn
CO
co
o
CD





O)
"O
X
o

S-
o
o
re
-t->
C
CD
O
cn
co

ID
o
CD
o


CO
in
ID
CO
ID
o
0







N

o
S-
o
f-
n3
^^
QJ
r— *


CO
o
o
CD


^
1 — 1
CM
!c
CD

CO
c
CD
;S
c
O)
o
1 —

o
o
s~
o
c_
rO
X
d)
CD
CM
i — i

T— 1
i-H
o
o


I — I
ID
cn
CM
cn
o
CD









s_
o
-C
u
o
-C
+^


1 — 1
1 — 1

CM
i — i
O
o



in
o
o










o

o



CD
in
cn

o
o
CD


^
o
cn
CO
o
o










c
r-
£
S-

2:
i — i
i — i
i — i

cn
CD
o
o


CO
CO
co
r-
0
o








c
S-

CLJ
E
c
d.

VI
(J

1 — 1

co
CD
O
O


CM
in
CO
CM
CO
O
o







•r-
~
4J
O>
c
n
oo

r
•^

i — i

i — t
o
CD



CM
O
CO
cn
o
o










i-

t.


Q-
O
p-H
ID

O
O
CD


CO
CO
1 —
UD
1 —
CD
O











O)
C
N

^~
CO
CO
CO
1 — 1

o
1 — 1
o
CD


^.
r-.
•>=(-
o
o










c
rd
S-
3


1—
                                               508.1-32

-------
 Figure 1. Aroclor 1016.  Chromatogram of LFB at 0.2 ug/L
                   20
                      25
                  TIME (MIN)
                                            30
                                     35
Figure 2. Aroclor 1221.  Chromatogram of LFB at 0.2 ug/L
        ' — 1
                       w
                  S
                  a.
                          i — i — I
                                                     JuJL^.
10
15
20
 25        30
TIME (MIN)
                                                 35
                                                 40
                              508.1-33

-------
Figure 3. Aroclor 1232.  Chromatogram of LFB at 0.2ug/L
                       v>
                                      V)

                                                                J	L.
10
15
20
  25         30
TIME (MIN)
                                                     35
                                                     40
Figure 4. Aroclor 1242.  Chromatogram of LFB at 0.2ug/L
    10
   15
  20
   25        30
 TIME(M1JN)
  508.1-34
                                                     35

-------
 Figures. Aroclor 1248.   Chroma tog rani of LFB at 0.2ug/L
           15
            20
            25        30
           TIME (MIN)
                                                    35
                                                     40
 Figure 6. Aroclor 1254.  Chromatogram of LFB at 0.2ug/L
                                            d   S
10
15
20
25        30
 TIME (MIN)
                                                 35
                                                  40
                                                 45
                                 508.1-35

-------
Figure 7.  Aroclor 1260.  Chromatogram of LFB at 0.2ug/L
            J	I
   10
  15
 20
 25        30
 TIME (MIN)
35
40
45
 Figure 8. Toxaphene.  Chromatogram of LFB at 0.2ug/L
                                                      •  .  I  •  i  •
10
15
20
25        30
 TIME (MIN)
 35
                                                         40
                                                         45
                                  508.1-36

-------
            METHOD  509.   DETERMINATION OF ETHYLENE THIOUREA (ETU)  IN
                         WATER USING  GAS  CHROMATOGRAPHY  WITH A
                         NITROGEN-PHOSPHORUS  DETECTOR
                                  Revision  1.1
                         Edited by J. W. Munch  (1995)
D.J. Munch (USEPA, Office of Water) and R.L. Graves (USEPA, NERL-Cincinnati)
T.M. Engel and S.T. Champagne, (Battelle, Columbus Division), National
Pesticide Survey Method 6, Revision 1.0, 1987

J.W. Eichelberger, Method 509 Revision 1.0 (1992)
                    NATIONAL EXPOSURE RESEARCH LABORATORY
                      OFFICE OF RESEARCH AND DEVELOPMENT
                     U.S.  ENVIRONMENTAL  PROTECTION AGENCY
                            CINCINNATI,  OHIO 45268
                                    509-1

-------
                                  METHOD 509

              DETERMINATION  OF  ETHYLENE THIOUREA  (ETU)  IN WATER
                       USING GAS CHROMATOGRAPHY WITH A
                         NITROGEN-PHOSPHORUS DETECTOR


1.   SCOPE AND APPLICATION

     1.1  This method utilizes gas chromatqgraphy (GC) to determine ethylene
          thiourea (ETU, Chemical Abstracts Registry No. 96-45-7) in water.

     1.2  This method has been validated in a single laboratory during
          development.  The method detection limit (MDL) has been determined
          in reagent water (1) and is listed in Table 2.  Method detection
          limits may vary among laboratories,  depending upon the analytical
          instrumentation used and the experience of the analyst. In addition
          to the work .done during the development of this method and its use
          in the National Pesticide Survey, an inter!aboratory method
          validation study of this method has been conducted.

     1.3  This method is restricted to use by or under the supervision of
          analysts experienced in the use of GC and in the interpretation of
          gas chromatograms.  Each analyst must demonstrate the ability to
          generate acceptable results with this method using the procedure
          described in Sect. 9.3.

     1.4  When a tentative identification of ETU is made using the recommended
          primary GC column  (Sect. 6.7.1), it must be confirmed by at least
          one additional qualitative technique.  This technique may be the use
          of the confirmation GC column (Sect. 6.7.2) with the nitrogen-
          phosphorus detector or analysis using a gas chromatograph/mass
          spectrometer  (GC/MS).

2.   SUMMARY OF METHOD

     2.1  The ionic strength and pH of a measured 50-mL aliquot of sample are
          adjusted by addition of ammonium chloride and potassium fluoride.
          The sample is  poured onto a column of kieselguhr diatomaceous earth.
          ETU is eluted  from the column with 400 ml of methylene chloride.  A
          free radical  scavenger is then added in excess to the eluate.  The
          methylene chloride eluant is concentrated to a volume of 5 ml after
          solvent exchange with ethyl acetate.  Gas chromatographic conditions
          are described  which permit the separation and measurement of ETU
          with a nitrogen-phosphorus detector  (NPD).

3.   DEFINITIONS

     3.1  ARTIFICIAL GROUND  WATER — An aqueous matrix designed to mimic a
          real ground water  sample.  The artificial ground water should be
          reproducible  for use by others.

                                     509-2

-------
 3.2  CALIBRATION STANDARD (CAL) — A solution prepared, from the primary
      dilution standard solution or stock standard solutions and the
      internal standards and surrogate analytes.   The CAL solutions are
      used to calibrate the instrument response with respect to analyte
      concentration.

 3.3  METHOD DETECTION LIMIT (MDL)  -- The minimum concentration of an
      analyte that can be identified, measured, and reported with 99%
      confidence that the analyte concentration is greater than zero.

 3.4  INTERNAL STANDARD (IS)  — A pure analyte(s)  added to a sample,
      extract, or standard solution in known amount(s)  and used to measure
      the relative responses  of other method analytes and surrogates  that
      are components of the same sample or solution.   The internal
      standard must be an analyte that is not a sample  component.

 3.5  FIELD DUPLICATES (FD1 and FD2)  — Two separate  samples collected  at
      the same time and place under identical  circumstances and treated
      exactly the same throughout field and laboratory  procedures.
      Analyses of FD1 and FD2 give  a  measure of the precision associated
      with sample collection,  preservation and storage,  as well  as  with
      laboratory procedures.

 3.6  INSTRUMENT PERFORMANCE  CHECK  SOLUTION (IPC)  -  A  solution  of  one  or
      more method analytes, surrogates,  internal  standards,  or other  test
      substances used to  evaluate the performance  of  the  instrument system
      with respect to a defined set of criteria.

 3.7  LABORATORY REAGENT  BLANK  (LRB)  -  An aliquot  of reagent water or
      other blank matrix  that  is  treated exactly as a sample including
      exposure to all  glassware,  equipment,  solvents, reagents,  internal
      standards,  and  surrogates that  are used  with  other  samples.   The  LRB
      is  used  to determine  if method  analytes  or other  interferences  are
      present  in the  laboratory environment, the reagents,  or the
      apparatus.

 3.8   QUALITY  CONTROL  SAMPLE  (QCS)  --  A  solution of method  analytes of
      known  concentrations which  is used  to  fortify an aliquot of LRB or
      sample matrix.   The QCS is  obtained  from  a source external to the
      laboratory  and different  from the  source  of calibration  standards.
      It  is  used  to check laboratory  performance with externally prepared
      test materials.

3.9   STOCK STANDARD SOLUTION (SSS) — A concentrated solution containing
      one  or more method analytes prepared  in the laboratory  using assayed
      reference materials or purchased from a reputable commercial source.


3.10 SURROGATE ANALYTE (SA) — A pure analyte(s), which is extremely
     unlikely to be found in any sample, and which is added to a sample
     aliquot  in known amounts(s) before extraction or other processing
     and  is measured with the  same procedures used to measure other

                                509-3

-------
          sample components.  The purpose of the SA is to monitor method
          performance with each sample.
4.   INTERFERENCES
     4.1
      4.2
      4.3
      4.4
      4.5
Method interferences from contaminants in solvents, reagents,
glassware and other sample processing apparatus may cause discrete
artifacts or elevated baselines in gas chromatograms.  All reagents
and apparatus must be routinely demonstrated to be free from
interferences under the conditions of the analysis by running
laboratory reagent blanks as described in Sect: 9.2.

4 1.1  Glassware must be scrupulously cleaned  (2).  Clean all glass-
       ware as soon as possible after use by thoroughly rinsing with
       the last solvent used in it.  Follow by washing with hot
       water and detergent and thorough rinsing with tap and reagent
       water.  Drain dry, and heat in an oven  or muffle furnace at
       400°C for 1 hr.  Do not heat volumetric ware.  Thermally
       stable materials might not be eliminated by this treatment.
       Thorough rinsing with acetone and methylene chloride may be
       substituted for the heating.  After drying  and cooling, seal
       and store glassware in a clean environment  to prevent any
       accumulation of dust or other contaminants.   Store inverted
       or capped with aluminum foil.

4.1.2  The use of high purity reagents and solvents  helps to
       minimize interference problems.   Purification of solvents  by
       distillation in all-glass  systems may be required.

Interfering contamination may occur when a sample  containing a low
concentration of ETU  is  analyzed  immediately following a  sample
containing a relatively  high concentration of  ETU.   Thorough
between-sample rinsing of the sample  syringe and  associated
equipment with ethyl  acetate can  minimize  sample  cross contamin-
ation.   After analysis of a  sample containing  high concentrations of
ETU,  one or more  injections  of ethyl  acetate should  be made  to
ensure that accurate  values  are  obtained for the  next  sample.

Matrix interferences  may be  caused by contaminants that  are
coextracted from  the  sample.  The extent of matrix interferences  may
vary considerably  from source to  source, depending upon  the  sample.
Tentative  identifications must  be confirmed using the  confirmation
column  (Sect. 6.7.2)  and the conditions  in Table  1.

Studies  have  shown  that  persistent  ETU decomposition is
circumstantially  linked  to  free  radical  mechanism.  Addition of  a
free radical  scavenger  is  necessary  to prohibit any free radical
reactions.

 It is important  that  samples and working standards be  contained  in
the same solvent.   The  solvent  for working standards must be the
 same as  the  final  solvent used  in sample preparation.   If this is
                                      509-4

-------
          not the case, chromatographic comparability of standards to sample
          may be affected.                                                >

5.   SAFETY

     5.1  ETU is a suspected carcinogen and teratogen.  Primary standards of
          ETU should be prepared in a hood.  A NIOSH/MESA approved toxic gas
          respirator should be worn when the analyst handles high concentra-
          tions of ETU.  Each laboratory is responsible for maintaining a
          current awareness file of OSHA regulations regarding the safe
          handling of the chemicals specified in this method.  A reference
          file of material data handling sheets should also be made available
          to all personnel involved in the chemical analysis.  Additional
          references to laboratory safety are available and have been
          identified (3-5) for the information of the analyst.

6.   EQUIPMENT AND SUPPLIES (All specifications are suggested.  Catalog
     numbers are included for illustration only.)

     6.1  SAMPLING CONTAINERS —  60-mL screw cap vials equipped with Teflon-
          faced silicone septa.  Prior to use, wash vials and septa with
          detergent and'rinse with tap and distilled water.  Allow the septa
          to air dry at room temperature, place in a 105°C oven for 1 hr, then
          remove and allow to cool in an area known to be free of organics.
          Heat vials at 400°C for 1 hr to remove organics.

     6.2  GLASSWARE

          6.2.1  Concentrator tube, Kuderna-Danish (K-D) - 10-mL or 25-mL,
                 graduated.  Calibration must be checked at the volumes
                 employed in the test.  Ground glass stoppers are used to
                 prevent evaporation of extracts.

          6.2.2  Evaporative flask, K-D - 500-mL Attach to concentrator tube
                 with springs.

          6.2.3  Snyder column, K-D - three-ball  macro to which a condenser
                 can be connected to collect solvent.

          6.2.4  Vials - Glass, 5 to,10-mL capacity with Teflon lined screw
                 caps.

     6.3  Boiling stones - carborundum,  #12 granules,  heat at 400°C for 30 min
          prior to use.  Cool  and store in a desiccator.

     6.4  Water bath - Heated,  capable of temperature  control  (±2°C).  The
          bath should be used in a hood.

     6.5  Balance - Analytical, capable of accurately  weighing to the nearest
          0.0001 g.
                                    509-5

-------
     6.6  Tube heater - Capable of holding 8 K-D concentrator tubes and
          heating the mid-section of the tubes to 35-40°C while applying a
          nitrogen stream.                '

     6.7  GAS CHROMATOGRAPH - Analytical system complete with GC equipped with
          a nitrogen-phosphorus detector, split/splitless injector for
          capillary columns and all required accessories.  A data system is
          recommended for measuring peak areas. An autoinjector is recommended
          to improve precision of analyses.

          6.7.1  Primary column - DB-Wax or equivalent,10 m x 0.25 mm I.D.
                 bonded fused silica column, 0.25 urn film thickness.
                 Validation data presented in this method were obtained using
                 this column.  Alternative columns may be used provided equal
                 or better peak separation and peak shape are obtained.

          6.7.2  Confirmation column - DB-1701 or equivalent, 5 m x 0.25 mm
                 I.D. bonded fused silica column, 0.25 urn film thickness.

          6.7.3  Detector - Nitrogen-phosphorus (NPD).  This detector has
                 proven effective in the analysis of ETU in fortified reagent
                 and artificial ground waters.

7.   REAGENTS AND STANDARDS

     7.1  REAGENT WATER — Reagent water is defined as water in which an
          interference is not observed at the retention time for ETU at the
          method detection limit.  A Millipore Super-Q Water System or its
          equivalent may be used to generate reagent water.  Water that has
          been charcoal filtered may also be suitable.

     7.2  Methylene chloride, ethyl acetate — distilled-in-glass quality or
          equivalent.

     7.3  Nitrogen gas - high purity.

     7.4  Extrelut QE Extraction column - Obtained from EM Science (Catalog
          No. 902050-1) or equivalent.  Extrelut QE columns contain a
          specially modified form of large pore Kieselguhr with a granular
          structure.

     7.5  Ammonium chloride, granular, ACS grade — for pH and ionic strength
          adjustment of samples.

     7.6  Potassium fluoride, anhydrous, ACS grade — for ionic strength
          adjustment of sample.

     7.7  Dithiothreitol (DTT) (Cleland's reagent) - for use as a free-radical
          scavenger (available from Aldrich Chemical Co.).

          7.7.1  DTT in ethyl acetate, 1000 /ig/ml_ - May be prepared by adding
                 1 g DTT to a 1-L volumetric flask and diluting to volume with

                                     509-6

-------
             ethyl  Acetate.   Smaller amounts may be prepared if only a
             small  number of samples are to be analyzed.   Store at room
             temperature.

 7.8  Propylene thiourea (PTU)  - For use as a surrogate standard.
      Prepared from carbon disulfide and 1,2-diaminopropane using  the
      procedure published by Hardtmann,  et. al.  (Journal  of Medicinal
.    ,  Chemistry,  18(5),  447-453, 1975),  or purchase from  commercial
      sources.

 7.9  3,4,5,6-Tetrahydro-2-pyrimidinethiol  (THP)  -  >98% purity,  for  use as
      an  internal  standard (available from Aldrich  Chemical  Co.).

 7.10  STOCK STANDARD SOLUTION (0.10  fig/pi)  - The  stock standard  solution
      may be purchased  as a  certified solution  or prepared  from  pure
      standard  material  using the  following procedure:

      7.10.1 Prepare stock standard  solution by  accurately  weighing
             0.0010 g of pure ETU.   Dissolve the  ETU  in ethyl  acetate
             containing  1000 /ig/mL of DTT  and dilute  to volume in  a  10-mL •
             volumetric .flask.   Larger volumes may.be used  at  the
             convenience of  the  analyst.   If ETU  purity is  certified at
             96% or greater,  the weight  may be used without  correction  to
             calculate the concentration of the  stock standard.
             Commercially prepared stock standards may  be used at  any
             concentration if they are certified  by the manufacturer or by
             an independent  source.

      7.10.2  Transfer the stock  standard solution into  a Teflon  sealed
             screw  cap vial.  Store  at room temperature and  protect  from
             light.

      7.10.3  The stock standard  solution should be replaced  after  two
             weeks  or sooner  if  comparison  with laboratory control
             standards indicates a problem.

7.11  INTERNAL STANDARD  FORTIFYING SOLUTION  — Prepare  an internal
      standard fortifying  solution by  accurately weighing 0.0010 g of pure
     THP.   Dissolve  the  THP  in  ethyl  acetate containing 1000 /ig/mL  of DTT
      and dilute to volume in a  10-mL  volumetric flask.  Transfer  the
     solution to a Teflon sealed screw cap  bottle and  store at room
     temperature.  Addition  of  50 nl  of the  internal standard fortifying
     solution to 5 mL of  sample extract results in a final   internal
     standard concentration  of  1.0 /zg/mL.

7.12 SURROGATE STANDARD  FORTIFYING SOLUTION - Prepare  a surrogate
     standard fortifying solution by  accurately weighing 0.0010 g of pure
     PTU.  Dissolve the PTU  in ethyl  acetate containing 1000 fig/ml of DTT
     and dilute to volume in a  10-mL volumetric flask.  Transfer the
     solution to a Teflon sealed screw cap bottle and  store at room
     temperature.   Addition of 5 /zL of the surrogate standard fortifying
     solution to a sample prior to extraction results  in  a  surrogate

                                509-7

-------
                                           I
          standard concentration in the sample of 10 /ig/L and, assuming
          quantitative recovery of PTU, a  surrogate standard concentration in
          the final extract of 0.10 /zg/mL.

     7.13 INSTRUMENT PERFORMANCE CHECK SOLUTION - Prepare the instrument
          performance check solution by adding 10 pi of the ETU stock standard
          solution, 1.0 mL of the internal standard fortifying solution, and
          100 nL of the surrogate standard fortifying solution to a 100-mL
          volumetric flask and diluting to volume with ethyl acetate
          containing 1000 /jg/mL of DTT.  Transfer the solution to a Teflon
          sealed screw cap bottle and store at room temperature.

8.   SAMPLE COLLECTION. PRESERVATION. AND  STORAGE

     8.1  SAMPLE COLLECTION — Grab samples must be collected in 60-mL glass
          containers fitted with Teflon-liiied screw caps (Sect. 6.1).  Conven-
          tional sampling practices (6) should be followed; however, the
          bottle must not be prerinsed with sample before collection.

     8.2  SAMPLE STORAGE — The samples must be iced or refrigerated at 4°C
          and protected from light from the time of collection until
          extraction.  Samples should be extracted as soon as possible after
          collection to avoid possible degradation of ETU.  All samples must
          be extracted within 14 days of collection.  Extracts must be stored
          under refrigeration and protected from light.  Extracts must be
          analyzed within 28 days of extraction.

     8.3  SAMPLE PRESERVATION — ETU may chemically degrade in some samples
          even when the sample is refrigerated.  When this method was
          developed, mercuric chloride was used to ensure against biological
          degradation.  No suitable preservation reagent has been found other
          than mercuric chloride.  However, the use of mercuric chloride is
          not recommended due to its toxicity and potential harm to the
          environment.  Biological degradation may occur only rarely in
          samples with limited biological  activity such as finished drinking
          waters.

9.   QUALITY CONTROL

     9.1  Each laboratory using this method is required to operate a formal
          quality control  (QC) program.  The minimum requirements of this
          program consist of the following;  an initial demonstration of
          laboratory capability;  measurement of the surrogate compound in
          each sample; analysis of laboratory reagent blanks, laboratory
          fortified blanks, laboratory fortified matrix samples, and QC check
          standards. A MDL for ETU must also be determined.
                                           I '

     9.2  LABORATORY REAGENT BLANKS — Before processing any samples, the
          analyst must demonstrate that all glassware and reagent
          interferences are Bunder control.  This is accomplished by analyzing
          a laboratory reagent blank (LRB).  A LRB is a 50-mL aliquot of
          reagent water,  fortified with the internal standard and the

                                    509-8

-------
      surrogate compound,  that is analyzed according to Sect.  11  exactly
      as if it were a sample.   Each time a set of samples is  analyzed  or
      reagents are changed,  it must be demonstrated that the  laboratory
      reagent blank is free  of contamination that would prevent the
      determination of ETU at  the MDL.  All  interfering contaminants must
      be eliminated before sample analyses are started.

 9.3   Initial  Demonstration  of Capability.

      9.3.1   Select a representative fortified concentration  (about 10
             times MDL or  at a concentration in the middle  of  the
             calibration range established in  Section  10) for  ETU.
             Prepare a 4-7 replicate LFBs  containing ETU at the selected
             concentration,  and  analyze each  LFB according to procedures
             beginning in  Sect.  11.

      9.3.2   The  mean recovery value for these samples  must fall  in the
             range of ± 20%  of the  fortified amount.  The precision of
             these measurements,  expressed as  RSD,  must  be  20% or less.
             If the data meet  these  criteria,  performance is considered
             acceptable.   If acceptance  criteria  is  not  met, this
             procedure must  be repeated  using  fresh  replicate  samples
             until  satisfactory performance  has been demonstrated.

      9.3.3   To determine  the  MDL,  prepare a minimum of  7 LFBs  at a low
             concentration.  The  fortification  concentration in Table  2
            may  be used as  a  guide,  or  use  calibration  data obtained  in
            Section  10 to estimate  a  concentration  that will  produce  a
            peak with  a 3-5 times  signal  to noise  response.   Extract  and
            analyze  each  replicate  according  to  Sections 11 and  12.   It
             is recommended  that  these LFBs  be  prepared  and analyzed over
            a  period  of several  days, so  that  day to day variations are
            reflected  in  the  precision  of the measurements.   Calculate
            mean  recovery and standard  deviation for each  analyte.  Use
            the  standard deviation  and  the  equation given  in  Section  13
            to calculate the  MDL.

     9.3.4  The  initial demonstration of  capability is used primarily to
            preclude  a laboratory from  analyzing unknown samples  via a
            new,  unfamiliar method  prior  to obtaining some experience
            with  it.   It is expected that as laboratory personnel gain
            experience with this method the quality of data will  improve
            beyond those required here.

9.4  The analyst  is permitted  to modify GC  columns or GC conditions to
     improve the  separations,  identifications, or lower the cost  of
     measurement.  Each time a modification is made, the analyst  is
     required to repeat the procedure in Sect. 9.3.
                                509-9

-------
9.5  ASSESSING SURROGATE  RECOVERY

     9.5.1  All  samples and blanks must be fortified with the surrogate
            compound  according to Sect. 11.1 before extraction to monitor
            preparation and analysis of samples.

     9.5.2  Surrogate recovery must be evaluated for acceptance by
            determining whether the measured surrogate concentration
            (expressed as percent recovery) falls within the required
            recovery  limits.  Performance-based recovery criteria for PTU
            has  been  generated from single-laboratory results.  Measured
            recovery  of PTU must be between 70 and 130 percent.

     9.5.3  If the surrogate recovery for? a sample or blank is outside of
            the  required  surrogate recovery limits specified in Sect.
            9.5.2, the laboratory must take the following actions:

            (1)      Check calculations to make sure there are no errors.

            (2)      Check internal standard and surrogate standard
                     solutions for degradation, contamination, or other
                     obvious abnormalities.

            (3)      Check instrument performance.
                                              •
                     Reinject the extract if the above steps fail to
                     reveal the cause of the problem.  The problem must
                     be identified and corrected before continuing.
                     Reanalyzing the sample or blank, if possible, may be
                     the only way to'solve the problem.

9.6  ASSESSING THE INTERNAL STANDARD

     9.6.1  The  analyst is must monitor the internal  standard peak area
            in all samples and blanks during each analysis day.  The IS
            response for any sample chromatogram should not deviate from
            the  IS response of the most recent daily calibration check
            standard by more than 20%.

     9.6.2  If >20% deviation occurs with an individual extract, optimize
            instrument performance and inject a second aliquot of that
            extract.   If the reinjected aliquot produces an acceptable IS
            response,  report results for that injection.  If a deviation
            >30% is obtained for the reinjected extract, reanalyze the
            sample beginning with Sect.  11, provided the sample is still
            available. Otherwise, report results obtained from the
            reinjected extract,  but mark them as suspect.

     9.6.3  If consecutive samples fail  the IS response acceptance
            criteria,  immediately analyze a medium calibration check
            standard.   If the check standard provides a response for the
            IS within 20% of the predicted value,  then follow procedures

                               509-10

-------
          .  itemized in Sect. 9.6.2 for each sample failing the IS
            response criteria.  If the check standard provides a response
            which deviates more than 20% from the predicted value, then
           .the analyst must recalibrate.

9.7  ASSESSING LABORATORY PERFORMANCE

     9.7.1  The laboratory must analyze at least one laboratory fortified
            blank (LFB) per sample set.  The ETU fortifying concentration
            in the LFB should be 10 times the MDL or at a concentration
            near the middle of the calibration range demonstrated by the
            laboratory.  Calculate the percent recovery of the ETU.  If
            the recovery falls outside the control limits (see Sect.
            9.7.2), the system is judged out of control and the source of
            the problem must be identified and resolved before continuing
            analyses.

     9.7.2  Until sufficient LFB data become available, usually a minimum
            of 20 to 30 results, the laboratory should assess its
            performance against the control limits described in Sect.
            9.3.2.  When sufficient laboratory performance data become
            available,  develop control  limits from the mean percent
            recovery (R) and standard deviation (S) of the percent
            recovery.  These data are used to establish upper and lower
            control limits as follows:

                     Upper Control  Limit = R + 3S
                     Lower Control  Limit = R - 3S

            After five to ten new recovery measurements are made,  control
            limits should be recalculated using only the most recent 20
            to 30 data points.  Control limits must not exceed the fixed
            acceptance limits in Section 9.3.2.

9.8  Assessing Analyte Recovery - Laboratory Fortified Sample Matrix

     9.8.1  The laboratory must add a known concentration to a minimum of
            5% of the routine samples or one sample per set, whichever is
            greater.   The fortified concentration should not be less than
            the background concentration of the sample selected for
            fortification.  Ideally,  the concentration should be the same
            as that used for the laboratory fortified blank (Sect.
            9.3.1).  Over time,  samples from all  routine sample sources
            should be fortified.

     9.8.2  Calculate the percent recovery, P of the concentration for
            each analyte,  after correcting the analytical  result,  X,  from
            the fortified sample for the background concentration, b,
            measured  in the unfortified sample, i.e.,:

                     P  = 100 (X - b)  /  fortifying concentration,


                               509-11

-------
                 and compare these values recoveries listed in Table 2.  The
                 calculated value of P must fall in the range of  ± 25% of the
                 amount fortified.  If P exceeds this control limit the
                 results for that analyte in the unfortified matrix must be
                 listed as suspect due to matrix interference.

     9.9  ASSESSING INSTRUMENT PERFORMANCE — Instrument performance should be
          monitored on a daily basis by analyzing the instrument performance
          check solution (IPC).  The IPC solution contains compounds monitor
          instrument sensitivity and column performance.  The IPC components
          and performance criteria are listed in Table 4.  Inability to
          demonstrate acceptable instrument performance indicates the need for
          remedial action on the GC-NPD system.  A chromatogram from the
          analysis of the IPC is shown in Figure 1.  The sensitivity
          requirements are set according the MDL.  MDLs will vary somewhat in
          different laboratories according to instrument capabilities.  The
          laboratory should adjust the amount of ETU in the IPC based on the
          demonstrated sensitivity of the instrumentation used.

     9.10 QC Samples-  It is recommended that the laboratory periodically (at ,
          least quarterly), analyze one or more standard materials from an
          outside source to validate performance.

     9.11 ADDITIONAL QC — It is recommended that the laboratory adopt
          additional quality assurance practices for use with this method.
          The specific practices that are most productive depend upon the
          needs of the laboratory and the nature of the samples.

10.  CALIBRATION AND STANDARDIZATION

     10.1 Establish GC operating parameters equivalent to those indicated in
          Table 1.  Ensure that the gas chromatographic system is working
          properly by injecting the instrument performance check solution
          (Sect. 7.14) and checking for proper peak shapes, reasonable
          retention times,  and sufficient sensitivity.  The GC system is
          calibrated using the internal standard technique (Sect. 10.2).

     10.2 INTERNAL STANDARD CALIBRATION PROCEDURE — This approach requires
          the analyst to use at least one internal  standard compatible in
          analytical behavior to the compound of interest.  The analyst must
          further demonstrate that the measurement  of the internal standard is
          not affected by method or matrix interferences.  In developing this
          method,  THP (3,4,5,6-tetrahydro-2-|)yrimidinethiol) was found to be a
         , suitable internal standard.

          10.2.1 Prepare ETU calibration standards  at five concentration
                 levels by adding volumes of the ETU stock standard solution
                 to five volumetric flasks.   To each flask, add a known
                 constant amount of internal  standard and dilute to volume
                 with ethyl acetate containing 1000 /jg/mL of DTT.  One of the
                 standards  should be representative of an ETU concentration
                 near, but  above, the MDL.   The other concentrations should

                                    509-12

-------
                 correspond to the range of concentrations expected  in the
                 sample concentrates, or should define the working range of
                 the detector.

          10.2.2 Inject each calibration standard and tabulate the relative
                 response for ETU to the internal standard (RRa) using the
                 equation:
                          RR  =  A/A,
                                     is
                 where:   Aa  =  the peak area of ETU, and
                          Ajs =  the peak area of the internal  standard.

                          Generate  a calibration curve of RR  versus ETU
                          concentration  in the sample in
          10.2.3 The working calibration curve must be verified on each
                 working day by the measurement of a minimum of two
                 calibration check standards, one at the beginning and one at
                 the end of the analysis day.  These check standards should be
                 at two different concentration levels to verify the  .
                 calibration curve.  For extended periods of analysis (greater
                 than 8 hrs.), it is strongly recommended that check standards
                 be interspersed with samples at regular intervals during the
                 course of the analyses.  If the ETU response varies from the
                 predicted response by more than 20%, the test should be
                 repeated using a fresh calibration standard.  Alternatively,  ,
                 a new ETU calibration curve should be prepared.  Any sample
                 extracts analyzed since the last acceptable calibration check
                 should be considered suspect, and should be reanalyzed after
                 calibration is restored.  •
11.   PROCEDURE

     11.1 SAMPLE EXTRACTION
          11.1.1 Pipet a 50-mL aliquot of water sample into a sample bottle
                 (Sect. 6.1) containing 1.5 g of ammonium chloride and 25 g of
                 potassium fluoride.  Seal the bottle and shake vigorously
                 until salts are dissolved.  Fortify the sample with 5 /il_ of
                 the surrogate standard fortifying solution (Sect. 7.13).

          11.1.2 Pour contents of the bottle onto the Extrelut (sorbent)
                 column (Sect. 7.4).  Allow the column to stand undisturbed
                 for 15 min.

          11.1.3 Add 5 ml of 1000 /ig/mL DTT in ethyl  acetate to a K-D
                 concentrator tube equipped with a 500-mL flask.

          11.1.4 Add 400 ml of methylene chloride in  50-75 ml portions to the
                 Extrelut column and collect the eluant in the K-D apparatus
                 (Sect. 11.1.3).

                                    509-13

-------
11.2 EXTRACT CONCENTRATION

     11.2.1 Conduct the following work in a fume hood which is properly
            vented.  Add 1 or 2 boiling stones to the K-D apparatus and
            attach a macro Snyder column.  Prewet the Snyder column by
            adding about 1 mL of met.hylene chloride to the top.  Attach a
            condenser to the Snyder column to recover the methylene
            chloride as it escapes the column. Place the K-D apparatus in
            a 65-70°C water bath so that the K-D tube is partially
            immersed in the hot water, and the entire lower rounded
            surface of the flask is bathed with hot vapor.  When the
            apparent volume of liquid reaches 5 ml, remove the K-D
            apparatus and allow it to drain and cool for at least 10 min.

     11.2.2 Reduce the liquid volume in the K-D tube to approximately 1
            ml by placing the sample in a tube heater at 35-40°C under a
            stream of nitrogen.  The tube heater heats the solvent in the
            K-D tube at volume markings between 1 and 10 ml.

     11.2.3 Dilute sample to 5 ml with ethyl acetate; rinse walls of K-D
            tube while adding ethyl acetate.  Immediately fortify the
            sample with 50 ^L of internal standard fortifying solution
            (Sect. 7.12).  Agitate sample to disperse internal standard.
            Transfer sample to a GC vial and determine ETU by GC-NPD as
            described in Sect. 11.3.

11.3 GAS CHROMATOGRAPHY

     11.3.1 Table 1 summarizes the recbmmended GC operating conditions.
            Included in Table 1 are retention times observed using this
            method.  An example of the separations achieved using these
            conditions are shown in Figure 1.  Other GC columns or
            chromatographic conditions may be used if the requirements of
            Sect. 9.3 are met.

     11.3.2 Calibrate or verify the system calibration daily as described
            in Sect. 10.  The standards and extracts must be in ethyl
            acetate.

     11.3.3 Inject 2 /iL of the sample extract.  Record the resulting peak
            size in area units.

     11.3.4 The width of the retention time window used to make
            identifications should be based upon measurements of actual
            retention time variations of standards over the course of a
            day.  Three times the standard deviation of a retention time
            can be used to calculate a suggested window size for a
            compound.  However, the experience of the analyst should
            weigh heavily in the interpretation of chromatograms.

     11.3.5 Confirmatory techniques such as chromatography with a
            dissimilar column, or an alternate technique such as particle

                               509-14

-------
                  beam/HPLC/mass spectrometry  (EPA Method 553) may be used for
                  confirmation of ETU in extracts prepared by this method.  A
                  suggested confirmation column is described in Table 1.

 12.  DATA ANALYSIS AND CALCULATIONS

      12.1 Calculate the ETU concentration in the sample from the ETU relative
           response (RRa)  to the internal  standard using the multi-point
           calibration curve described in Sect. 10.2.2.   Do not use the daily
           calibration verification standard to quantitate ETU in samples.  Do
           not extrapolate beyond the linear range established during
           calibration.

 13.  METHOD PERFORMANCE

      13.1 .In a single 'laboratory,  ETU recovery and precision data from reagent
           water were determined at four concentration levels.   Results were
           used to determine the MDL and demonstrate method range.  These data
           are given in Table 2.  The equation used to calculate the MDL are as
           follows:
                           MDL   S t(lv1|1_alpha = 0_99)
                  where:
                             --    = o.99i = Student's t value for the 99%
                           confidence level  with n-1 degrees of freedom

                           n = number of replicates

  '..';'                S = standard deviation of replicate analyses.


      13.2  In  a single laboratory,  ETU recovery and precision  data from two
           artificial  ground waters were determined at a  single concentration
           level  of 10 /jg/L.  Results were used to demonstrate applicability of
           the method  to different  ground water matrices.   These  data are
           listed in Table 3.

,14.   POLLUTION PREVENTION

      14.1  Although this method requires 400  mL methylene  chloride extracting
           solvent  per sample,  no pollution  of  the environment will  occur due
           to  the recovery of the solvent during the extract concentration
           procedure.   Very little  solvent will  escape the  fume hood.   No other
           solvents are utilized in this method except for  the very  small
           amount of ethyl  acetate  needed to  make up calibration  and
           fortification standards.  These small  amounts of solvent  pose  no
           threat to the environment..-.

      14.2  For information  about pollution prevention that  may be applicable to
           laboratory  operations, consult "Less  is Better:   Laboratory Chemical

                                     509-15

-------
          Management for Waste Reduction" available from the American Chemical
          Society's Department of Government Relations and Science Policy,
          1155 16th Street N.W., Washington, D.C. 20036.

15.  WASTE MANAGEMENT

     15.1 It is the laboratory's responsibility to comply with all federal,
          state, and local regulations governing waste management,
          particularly the hazardous waste identification rules, and land
          disposal restrictions.  The laboratory has the responsibility to
          protect the air, water, and land by minimizing and controlling all
          releases from fume hoods and bench operations.  Compliance is also
          required with any sewage discharge permits and regulations.  For
          further information on waste management, consult "The Waste
          Management Manual for Laboratory Personnel," also available from the
          American Chemical Society at the address in Sect. 14.2.

16.  REFERENCES

     1.   40 CFR, Part 136, Appendix B

     2.   ASTM Annual Book of Standards, Pa,rt 31, D3694, "Standard Practice
          for Preparation of Sample Containers and for Preservation," American
          Society for Testing and Materials, Philadelphia, PA, p. 679, 1980.

     3.   "Carcinogens - Working with Carcinogens," Department of Health,
          Education, and Welfare, Public Health Service, Center for Disease
          Control, National Institute for Occupational Safety and Health,
          Publication No. 77-206, Aug. 1977,

     4.   "OSHA Safety and Health Standards, General Industry," (29 CFR 1910),
          Occupational Safety and Health Administration, OSHA 2206, (Revised,
          January 1976).

     5.   "Safety in Academic Chemistry Laboratories," American Chemical
          Society Publication, Committee on Chemical Safety, 3rd Edition,
          1979.

     6.   ASTM Annual Book of Standards, Part 31, D3370, "Standard Practice
          for Sampling Water,"  American Society for Testing and Materials,
          Philadelphia, PA, p. 76, 1980.
                                    509-16

-------
 17.   TABLES.  DIAGRAMS.  FLOWCHARTS.  AND VALIDATION DATA
         TABLE 1.  PRIMARY AND CONFIRMATION CHROMATOGRAPHIC CONDITIONS
Analyte
                                             Retention  Time,  min
      Primary column
Confirmation column
ETU
THP  (internal  standard)
PTU  (surrogate  standard)
           3.5
           5.1
           2.7
        4.5
        5.0
        2.2
     Primary conditions:

                 Column:

            Carrier gas:
             Makeup gas:
         Detector gases:
   Injector temperature:
   Detector temperature:
       Oven temperature:
                 Sample:
               Detector:

Confirmation conditions:

                 Column:

            Carrier gas:
             Makeup gas:
         Detector gases:
   Injector temperature:
   Detector temperature:
       Oven temperature:
                 Sample:
               Detector:
 10 m  long x 0.25 mm  I.D. DB-Wax bonded  fused
 silica column  (J&W), 0.25 m film thickness
 He @  30 cm/sec  linear velocity
 He @  30 mL/min  flow
 Air @ 100 mL/min flow; H2 @ 3 mL/min flow
 220°C
 230°C
 220'°C isothermal
 2 ill  splitless; 9 sec split delay
 Nitrogen-phosphorus
5 m long x 0.25 mm I.D. DB-1701 bonded fused
silica column (J&W), 0.25 m film thickness
He @ 30 cm/sec linear velocity
He @ 30 mL/min flow
Air G> 100 m:/min flow; H2 @ 3 mL/min flow
150°C
270°C
150°C isothermal
2 (J.L splitless;  9 sec split delay
Nitrogen-phosphorus
                                    509-17

-------
           TABLE 2.  RESULTS FROM MDL AND METHOD RANGE STUDIES (a)
Fortified
Level ,
5.0
10
25
100
Amt in
Blank,
0.492
ND (b)
ND
ND

n(d)
7
7
7
7
.. • - . ! - '
R(e)
97 (c)
102
94
97

S(f)
0.845
0.886
1.31
5.96

RSD(g)
17
9
6
6

MDL
2.7
-

(a)    Studies conducted in reagent water; average recovery of PTU surrogate
      from seven fortified reagent water samples was 100% (RSD) was 8.5/«.
(b)    ND = not detected.
(c)    Data corrected for amount detected in blank.
(d)    n = number of recovery data points.
(e)    R ~ aveVage percent recovery.
(f)    S = standard deviation.
(g)    RSD = percent relative standard deviation.
                                     509-18

-------
              TABLE 3.   RESULTS FROM MATRIX EVALUATION STUDIES (a)
Matrix
Hard (b)
Organic-contaminated (c)
(a) Samples were fortified
Amt. in
Blank,
M9/L
ND (d)
ND
at 10 fj.g/1 level
(b) Absopure Natural Artesian Spring water
n(e) R(f) .
7 93
7 93
with ETU.
obtained from th
S(g) RSD(h)
0.372 4
0.253 3

IB Absooure Water
      Company in Plymouth, Michigan.

(c)   Reagent water fortified with fulvic acid at the 1 mg/L concentration
      level.  A well-characterized fulvic acid, available from the
      International Humic Substances Society (associated with the United
      States Geological Survey in Denver, Colorado), was used.

(d)   ND = not detected.

(e)   n = number of recovery data points.

(f)   R = average percent recovery.

(g)   S = standard deviation.

(h)   RSD = percent relative standard deviation.
                                    509-19

-------
                         0)
                         CU
                         CL

                         cu














*^
CJ
UJ
3=
O

UJ
t *
s^
^c
V-
t*t*-
o
u_
o:
UJ
o.
>.
OS
O
H-
*5
o
CD
3


«a-
LU
_J
CQ
«t
L_»






.

































rO
^-*
00
4J
c
cu
pi
cu
i-
• r~>
3
cr
cu
cr:









_ _j
0 E
c ~-~
o en
0 =1












cu
+->
^
ns
C
^^
























4.)
oo
cu

CO
A
^
^^
OO

cu
4-3
>^
F—
ns
£=
ns

(4—
O

c
o

4-3
0
cu
4-3
cu
Q




1—4
o

0




l*~**
l^
UJ

nS
CU

3
O
••—
^

CU

cu
r_
>^
-C
^J
LU

.









>j
4_9
•r-
>•
•r—
•1— ^
•f—
00
C
CU

--~.^->
«2.^.
LO l-~
0 0

• — 1 1 — i
T3 -O
C £=
nS nS

LO CO
cn crt
• •
0 0

c c:
cu cu
cu cu
3 3
+J •»-»
cu cu
-Q -Q

LJ_ U_
OO CD
Q. Q_




i — i






CL.
^^
t—

o 'o
S- -r-
^3 <""
>, 4-5
J= CU
OJ C
S- -r-
CU -i-
H- E
1 -•-
10 S-
*•> >>
LD Q.
« 1
^J- CM
n
CO







o
.r- OJ
-c o
a. c:
ns ns
s_ e
en S-
o o
4-3 ^4—
ns S_
E CU
O CL
s_
5
1





















•
c
o
•r-
4_3
n)
^
CT
CU
CU

4 ^

CO
.^
oo
3
•o
cu
ns
^
O

nS
f ^


S-
O
4->
0
ns
C4»

^•^
S-

cu
e
1 E **""•*
>j ^\N
t/1 **~*
^
™^
1 ffl
CD
Q-

II II

LLa Ll_
00 OO
CL CL.



i^«^
•-2-
00
•r—
^-,
iS
'3

-o
ns

•
oo
o
cu
00

£=


j \
f—
CO
•f—
cu
-C

c^.
^
nJ


+J
ns
\x
ns
CU
CL
cu
.c:

ti_
0
4->
0
S-


cu
.c

t4_
o

.c

•o •
•r- 4J
3 f

^-> CU •!-
-\~ -CO)

™g
00 <4-
X -i-i—
(tj
m ^-^.c
• ^yy
O — **->
^B ^9

eu •
S- 00
cu u
-C CU
3 oo



























• •
C
o

I ^
ns
3
cr
cu
cu

4J

en
.^
oo
3
•^
cu
. to
3
O

nS
t ^


jj
O

u
ns


C
ns

OO
00
3
ns
f ^

N^
A
CU
CL

II

U_
CJJ
CL.



^~^
S

























































^^
ptY*1
*^_^
3

X

co
oo

1—4


II

U-
o
CL.





509-20

-------
                 V  ri
                    k*
                      R
                      4
                    Ll
                              s
509-21

-------
THIS PAGE LEFT BLANK INTENTIONALLY
              509-22

-------
     METHOD  515.1.   DETERMINATION OF CHLORINATED  ACIDS  IN  WATER BY  GAS
                     CHROMATOGRAPHY WITH AN  ELECTRON  CAPTURE  DETECTOR
                                 Revision 4.1


                          Edited by J.W. Munch (1995)




R.C. Dressman and J.J. Lichtenberg - EPA 600/4-81-053, Revision 1.0  (1981)

J.W. Hodgeson - Method 515, Revision 2.0 (1986)

D. J. Munch (USEPA, Office of Water) and T. Engel (Battelle Columbus
            Laboratories) - National Pesticide Survey Method 3, Revision 3 0
            (1987)

R.L. Graves - Method 515.1, Revision 4.0 (1989)
                    NATIONAL EXPOSURE RESEARCH LABORATORY
                      OFFICE OF RESEARCH AND DEVELOPMENT
                     U.S.  ENVIRONMENTAL  PROTECTION AGENCY
                           CINCINNATI, OHIO  45268
                                   515.1-1

-------
                                METHOD 515.1

              DETERMINATION OF CHLORINATED ACIDS IN WATER BY GAS
               CHROMATOGRAPHY WITH AN ELECTRON CAPTURE DETECTOR
1.   SCOPE AND APPLICATION
     1.1
      1.2
      1.3
This is a gas chromatographic (GC) method applicable to the
determination of certain chlorinated acids in ground water and
finished drinking water.  The following compounds can be determined
by this method:
          Analvte

          Acifluorfen*
          Bentazon
          Chloramben*
          2,4-D
          Dalapon*
          2,4-DB
          DCPA acid metabolites(a)
          Dicamba
          3,5-Dichlorobenzoic acid
          Dichlorprop
          Dinoseb
          5-Hydroxydi camba
          4-Nitrophenol*
          Pentachlorophenol  (PCP)
          Picloram
          2,4,5-T
          2,4,5-TP
                                  Chemical Abstract Services
                                        Registry Number

                                          50594-66-6
                                          25057-89-0
                                            133-90-4
                                             94-75-7
                                             75-99-0
                                             94-82-6
                                           1918-
                                             51-
                                            120-
                                             88-
                                           7600
                                            100
                                             87
                                           1918
                                             93
                                             93
-00-9
•36-5
-36-5
-85-7
-50-2
-02-7
-86-5
-02-1
-76-5
-72-1
 (a)DCPA monoacid  and  diacid  metabolites  included  in method  scope;
 DCPA diacid  metabolite  used  for  validation  studies.

 *These compounds  are  only  qualitatively  identified.  These  compounds
 are not quantitated because  control  over precision has  not  been
 accomplished.

 This method  is  also applicable to  the  determination of  salts  and
 esters of analyte acids.   The form of  each  acid  is not  distinguished
 by this method.   Results  are calculated  and reported for  each listed
 analyte as the  total  free  acid.

 This method  has been  validated  ijfi-a single  laboratory  and estimated
 detection limits  (EDLs) and  method detection limits  (MDLs)  have  been
 determined for  the analytes  above  (Sect.13).  Observed  detection
 limits may vary between ground waters,  depending upon  the nature  of
 interferences  in  the  sample  matrix and the  specific  instrumentation
 used.
                                    515.1-2

-------
      1.4  This method is restricted to use by or under the supervision of
           analysts experienced in the use of GC and in the interpretation of
           gas chromatograms.  Each analyst must demonstrate the ability to

                                                              the
      1.5
     Analytes that are not separated chromatographically i.e., which have
     very similar retention times, cannot be individually identified and
     measured in the same calibration mixture or water sample unless an
     alternate technique for identification and quantitation exist (Sect.
     •I J. • -7 ) •

1.6  When this method is used to analyze unfamiliar samples for any or
     all  of the analytes above,  analyte identifications must be confirmed
     by at  least one additional  qualitative technique.
 2.    SUMMARY OF METHOD
      2.1   A measured-volume of sample of approximately li is  adjusted  to  PH
           12 with  6  N sodium hydroxide and  shaken  for 1  hr to  hydrolyze
           derivatives.  (  Note:  Since  many of the herbicides contained  in this
           method are applied as a  variety of esters  and  salts,  it  is vital  to
           hydrolyze  them  to the parent acid prior  to extraction.)   Extraneous
           organic  material  is  removed by a  solvent wash.   The  sample is acidi-
           fied, and  the chlorinated acids are  extracted  with ethyl  ether by
           shaking  in a separatory  funnel  or mechanical tumbling  in  a bottle
           The acids  are converted  to  their  methyl  esters using diazomethane'as
                                             '
           ™nc                              , trimethylsilyldiazomethane
           (IMSD).  Excess denvatizing reagent  is removed, and the esters are
          determined by capillary column/GC using an electron capture detector
           (hLL) ) .
                                                            i

     2.2  The method provides aa optional Florisil separation procedure to aid
          in the elimination of interferences that may be encountered.

3.   DEFINITIONS

     3.1  INTERNAL STANDARD - A pure analyte(s) added to a solution in known
          amount (s) and used to measure the relative responses of other method
          analytes and surrogates that are components of the same solution
          The internal  standard must be an analyte that is not a sample
          component.                                                 K

     3.2  SURROGATE ANALYTE - A pure analyte(s),  which is extremely unlikely
          to be  found  in any sample,  and  which is  added to a sample  aliquot  in
          known  amount (s)  before extraction  and  is measured  with  the same
          procedures  used  to measure  other sample /components.   The purpose of
          a surrogate  analyte is to  monitor  method performance with  each
          sample.

     3.3   LABORATORY DUPLICATES  (LD1  and  LD2)  - Two  sample  aliquots  taken in
          the analytical laboratory  and analyzed separately with  identical
          procedures.   Analyses  of LD1  and LD2 give a measure  of  the  precision


                                   515.1-3

-------
     associated with laboratory procedures, but not with sample
     collection, preservation, or storage procedures.

3.4  FIELD DUPLICATES (FD1 and FD2) — Two separate samples collected at
     the same time and place under identical  circumstances and treated
     exactly the same throughout field and laboratory procedures.
     Analyses of FD1 and FD2 give a measure of the precision associated
     with sample collection, preservation and storage, as weTl as  with
     laboratory procedures.

3.5  LABORATORY REAGENT BLANK (LRB) — An aliquot of reagent water that
     is treated exactly as a sample including exposure to all glassware,
     equipment, solvents, reagents, internal  standards, and surrogates
     that are used with other samples.  The LRB is used to determine if
     method analytes or other interferences are present in the laboratory
     environment, the reagents, or the apparatus.                    ,

3.6  FIELD REAGENT BLANK (FRB) — Reagent water placed in a sample
     container  in the laboratory and treated as a sample in all respects,
     including  exposure to sampling site conditions, storage,
     preservation and all analytical procedures. The purpose of the FRB
     is to determine if method analytes or other interferences are
     present in the field environment.

3.7  LABORATORY PERFORMANCE CHECK SOLUTION (LPC) — A solution of method
     analytes,  surrogate compounds, and internal standards used to
     evaluate the performance of the instrument system with respect to a
     defined set of method criteria.     |

3.8  LABORATORY FORTIFIED BLANK  (LFB) — An aliquot of reagent water to
     which known quantities of the method  analytes are added  in the
     laboratory.  The LFB is  analyzed exactly  like a  sample,  and its
     purpose is to determine  whether the methodology  is  in control, and
     whether the laboratory is capable of  making accurate  and  precise
     measurements at the required method detection limit.

3.9  LABORATORY FORTIFIED SAMPLE MATRIX  (jLFM)  — An aliquot of an
     environmental sample to  which  known quantities of the method
     analytes  are added'in  the laboratory.  The  LFM is analyzed exactly
     like a  sample, and  its purpose is to  determine whether the sample
     matrix  contributes  bias  to  the analytical  results.  The  background
     concentrations of the  analytes in the sample  matrix must  be
     determined in a separate aliquot and  the  measured values  in the  LFM
     corrected  for background concentrations.

3.10 STOCK STANDARD SOLUTION  —  A  concentrated solution  containing  a
     single  certified standard that is a method  analyte,  or  a
     concentrated solution  of a  single analyte prepared  in  the laboratory
     with an assayed reference compound.   Stock standard  solutions  are
     used to prepare primary  dilution standards.

3.11 PRIMARY DILUTION STANDARD SOLUTION  — A  solution' of several  analytes
     prepared  in  the laboratory  from  stock standard  solutions and  diluted


                               515.1-4

-------
           solutions/0 ^^^ Ca1ibration S°lut1ons  and other needed  analyte


      3.12  CALIBRATION  STANDARD (CAL)  -  A solution  prepared  from  the primary
           dilution  standard  solution  and stock  standard  solutions  of the
           McIlTJ    standards  and  surrogate analytes.   The  CAL  solutions are
           used to calibrate  the instrument response with respect  to analvte
           concentration.                                            a.ia.jrLe

     3.13  QUALITY CONTROL SAMPLE  (QCS) --  A sample matrix  containing method
           analytes  or  a solution  of method analytes in a water miscible
           solvent which is used to fortify reagent water or environmental
           samples.  The QCS is obtained  from a  source external to the
           laboratory,  and is used to check laboratory performance with
           externally prepared test materials.

4.   INTERFERENCES

     4.1  Method interferences may be caused by contaminants in solvents
          reagents  glassware and other sample processing apparatus that'lead
          to discrete artifacts or elevated baselines in gas chromatograms.
          All  reagents  and apparatus must be routinely demonstrated to  be free
          from interferences under the conditions of the analysis by running
          laboratory reagent blanks as described in  Sect. 9.2.

          4.1.1   Glassware must be scrupulously  cleaned.(1)  Clean all  glass-
              .   ware as soon as possible after  use  by thoroughly rinsing  with
                 the last  solvent  used in it.   Follow by washing with hot
                 water  and detergent  and  thorough rinsing with  dilute acid
                 tap and reagent water.   Drain dry,  and heat  in an oven or'
                 muffle furnace at 400°C  for  1 hr.   Do not  heat volumetric
                 glassware.   Thermally stable materials  such  as PCBs might not
                 be  eliminated  by  this  treatment.  Thorough rinsing with
                 acetone may  be substituted for  the  heating.  After drying and
                 cooling,  seal  and store  glassware in  a clean environment  to
                 prevent any  accumulation  of dust or other  contaminants
                 Store  inverted  or capped  with aluminum foil.

         4.1.2  The use of high purity reagents  and solvents helps to
                minimize  interference problems.  Purification  of  solvents by
                distillation in all-glass  systems may be required
                WARNING:  When a  solvent  is purified, stabilizers added by
                the manufacturer  are removed, thus potentially making the
                solvent hazardous.  Also,  when  a solvent is purified
                preservatives added by the manufacturer are removed 'thus
                potentially reducing the shelf-life.               '

    4.2  The acid forms of the analytes are strong organic acids which  react
         readily with alkaline substances  and can be lost during sample
         preparation.  Glassware and glass wool  must be acid-rinsed with IN
         hydrochloric acid and the sodium sulfate must be acidified with
         sulfuric acid  prior to use to avoid analyte losses due to
         adsorption.


                                   515.1-5

-------
     4.3  Organic acids and phenols, especially chlorinated compounds,  cause
          the most direct interference with the determination.   Alkaline
          hydrolysis and subsequent extraction of the basic sample removes
          many chlorinated hydrocarbons and phthalate esters that might
          otherwise interfere with the electron capture analysis.

     4.4  Interferences by phthalate esters can pose a major problem in pesti-
          cide analysis when using the ECD.  These compounds generally  appear
          in the chromatogram as large peaks.  Common flexible  plastics
          contain varying amounts of phthalates, that are easily extracted or
          leached during laboratory operations.  Cross contamination of clean
          glassware routinely occurs when plastics are handled  during
          extraction steps, especially when solvent-wetted surfaces are
          handled.  Interferences from phthalates can best be minimized by
          avoiding the use of plastics in the laboratory.  Exhaustive
          purification of reagents and glassware may be required to eliminate
          background phthalate contamination.(1)

     4.5  Interfering contamination may occur when a sample containing  low
          concentrations of analytes is analyzed immediately following  a
          sample containing relatively high concentrations of analytes.
          Between-sample rinsing of the sample syringe and associated
          equipment with methyl-t-butyl-ether (MTBE) can minimize sample cross
          contamination.  After analysis of a sample containing high
          concentrations of analytes, one or more injections of MTBE should be
          made to ensure that accurate values are obtained for the next
          sample.

     4.6  Matrix interferences may be caused by contaminants that are
          coextracted from the sample.  Also, note that all analytes listed in
          the Scope and Application Section are not resolved from each  other
          on any one column, i.e., one analyte of interest may be an
          interferant for another analyte of interest.  The extent of matrix
          interferences will vary considerably from source to source,
          depending upon the water sampled.  The procedures in Sect. 11 can be
          used to overcome many of these interferences.  Positive
          identifications should be confirmed (Sect. 11.9).

     4.7  It is important that samples and working standards be contained in
          the same solvent.  The solvent for working standards must be  the
          same as the final solvent used in sample preparation.  If this is
          not the case, chromatographic comparability of standards to sample
          may be affected.
5.   SAFETY
     5.1  The toxicity or carcinogenicity of each reagent used in this method
          has not been precisely defined; however, each chemical compound'must
          be treated as a potential health hazard.  Accordingly, exposure to
          these chemicals must be reduced; to the lowest possible level.  The
          laboratory is responsible for maintaining a current awareness file
          of OSHA regulations regarding the safe handling of the chemicals
          specified in this method.  A reference file of material safety data
          sheets should also be made available to all personnel involved in

                                    515.1-6

-------
      the  chemical  analysis.   Additional  references  to  laboratory safety
      are  available  and  have  been  identified  (2-4) for  the  information of
      the  analyst.                      '

 5.2  DIAZOMETHANE — A  toxic  carcinogen  which can explode  under certain
      conditions.  The following precautions must be followed:

      5.2.1  Use only a  well ventilated hood — do not  breath vapors.

      5.2,2  Use a safety screen.

      5.2.3  Use mechanical pipetting aides.

      5.2.4  Do not heat above 90°C — EXPLOSION may result.

      5.2.5  Avoid grinding surfaces, ground glass joints, sleeve
             bearings,  glass stirrers — EXPLOSION may result.

      5.2.6  Store away from alkali metals — EXPLOSION may result.

      5.2.7  Solutions  of diazomethane decompose rapidly in the presence
             of solid materials such as copper powder,  calcium chloride
             and boiling  chips.                                        '

      5.2.8  The diazomethane  generation apparatus used in the
             esterification  procedures (Sect.  11.4 and  11.5) produces
             micromolar  amounts of  diazomethane  to minimize safety
             hazards.

 5.3   ETHYL ETHER —  Nanograde,  redistilled  in glass,  if necessary.

      5.3.1  Ethyl  ether  is  an  extremely flammable solvent.   If  a
             mechanical device  is used for sample  extraction,  the device
             should  be equipped  with  an  explosion-proof  motor  and placed
             in  a  hood to avoid  possible damage and injury  due to an
             explosion.

      5.3.2  Ethyl ether  must  be free  of peroxides as indicated  by EM
             Quant test strips  (available  from Scientific  Products Co
             Cat. No. PI 126-8,  and  other suppliers).

5.4  WARNING:  When a solvent  is purified, stabilizers  added by the
     manufacturer are removed, thus potentially making  the  solvent
     hazardous.

EQUIPMENT AND SUPPLIES (All specifications are suggested.  Catalog
numbers  are included for illustration only.)

6.1  SAMPLE BOTTLE — Borosil icate, 1-L volume with graduations (Wheaton
     Media/Lab bottle 219820 or equivalent),  fitted with screw caps  lined
     with TFE-fluorocarbon.   Protect samples from light.  The container
     must be washed and dried as described in Sect.  4.1.1 before use to
     minimize contamination.  Cap liners are cut to fit from sheets


                               515.1-7

-------
     (Pierce Catalog No. 012736) and extracted with methanol  overnight
     prior to use.

6.2  GLASSWARE

     6.2.1  Separatory funnel — 2000-mL, ;with TFE-fluorocarbon stop-
            cocks, ground glass or TFE-flilorocarbon stoppers.

     6.2.2  Tumbler bottle — 1.7-L (Wheaton Roller Culture Vessel or
            equivalent), with TFE-fluorocarbon lined screw cap.  Cap
            liners are cut to fit from sheets (Pierce Catalog No. 012736)
            and extracted with methanol overnight prior to use.

     6.2.3  Concentrator tube, Kuderna-Danish (K-D) — 10- or 25-mL,
            graduated (Kontes K-570050-2525 or Kontes K-570050-1025 or
            equivalent).  Calibration must be checked at the volumes
            employed in the test.  Ground ;glass stoppers are used to
            prevent evaporation of extracts.

     6.2.4  Evaporative flask, K-D — 500-mL (Kontes K-570001-0500 or
            equivalent).  Attach to concentrator tube with springs.

     6.2.5  Snyder column, K-D — three-ball macro  (Kontes K-503000-0121
            or equivalent).

     6.2.6  Snyder column, K-D — two-ball micro (Kontes K-569001-0219 or
            equivalent).

     6.2.7  Flask, round-bottom — 500-mL with 24/40 ground  glass joint.

     6.2.8  Vials — glass, 5- to 10-mL  capacity with TFE-fluorocarbon
            lined screw cap.

     6.2.9  Disposable pipets — sterile  plugged borosilicate glass, 5-mL
            capacity (Corning 7078-5N or  equivalent).

6.3  SEPARATORY FUNNEL SHAKER — Capable  of holding 2-L separatory
     funnels and  shaking them with rocking motion  to achieve thorough
     mixing of separatory funnel contents  (available from  Eberbach Co. in
     Ann Arbor, MI or other  suppliers).

6.4  TUMBLER — Capable of holding tumbler bottles  and tumbling  them
     end-over-end at 30 turns/min  (Associated Design and Mfg. Co.,
     Alexandria,  VA and other suppliers).

6.5  BOILING STONES — Teflon,  Chemware  (Norton  Performance  Plastics No.
     015021 and other suppliers).

6.6  WATER  BATH — Heated, capable of  temperature  control  (± 2°C).  The
     bath should  be used  in  a hood.

6.7  BALANCE — Analytical,  capable of accurately  weighing to  the nearest
     0.0001 g.
                               515.1-8

-------
6.8  DIAZOMETHANE GENERATOR — Assemble from two 20 x 150 mm test tubes,
     two Neoprene rubber stoppers, and a source of nitrogen as shown in
     Figure 1  (available from Aldrich Chemical Co.)-  When esterification
     is performed using diazomethane solution, the diazomethane collector
     is cooled in an approximately 2-L thermos for ice bath or a
     cryogenically cooled vessel  (Thermoelectrics Unlimited Model SK-12
     or equivalent).                                   .

6.9  GLASS WOOL — Acid washed (Supelco 2-0383 or equivalent) and heated
     at 450°C  for 4 hr.

6.10 GAS CHROMATOGRAPH — Analytical system complete with temperature
     programmable GC suitable for use with capillary columns and all
     required  accessories including syringes, analytical columns, gases,
     detector  and stripchart recorder or computerized data system.  A
     data system is recommended for measuring peak areas.  Table 1 lists ,
     retention times observed for method analytes using the columns and
     analytical conditions described below.

     6.10.1 Column 1 (Primary column) — 30 m long x 0.25 mm I.D. DB-5
            bonded fused silica column, 0.25 fim film thickness (J&W
            Scientific).  Helium carrier gas flow is established at 30
            cm/sec linear velocity and oven temperature is programmed
            from 60°C to 300°C at 4°C/min.  Data presented in this method
            were obtained using this column.  The injection volume was
            2 ill splitless mode with 45 second delay.  The injector
            temperature was 250°C and the detector was 320°C.  Alterna-
            tive columns may be used in accordance with the provisions
            described in Sect. 9.4.

     6.10.2 Column 2 (Confirmation column) — 30 m long x 0.25 mm I.D.
            DB-1701 bonded fused silica column, 0.25 /im film thickness
            (J&W Scientific).  Helium carrier gas flow is established at
            30 cm/sec linear velocity and oven temperature is programmed
            from 60°C to 300°C at 4°C/min.

     6.10.3 Detector — Electron capture.  This detector has proven
            effective in the analysis of method analytes in fortified
            reagent and artificial ground waters.

REAGENTS AND STANDARDS - WARNING:  When a solvent is purified,
stabilizers added by the manufacturer are removed, thus potentially
making the solvent hazardous.  Also, when a solvent is purified,
preservatives added by the manufacturer are removed, thus potentially
reducing the shelf-life.

7.1  ACETONE, METHANOL, METHYLENE CHLORIDE, MTBE — Pesticide quality or
     equivalent.

7.2  ETHYL ETHER, UNPRESERVED — Nanograde, redistilled in glass if
     necessary.  Must be free of peroxides as indicated by EM Quant test
     strips (available from Scientific Products Co., Cat. No. PI126-8,
     and other suppliers).   Procedures recommended for removal of per-
     oxides are provided with the test strips.

                               515.1-9

-------
 7.3   SODIUM SULFATE,  GRANULAR,  ANHYDROUS,  ACS  GRADE  —  Heat  treat  in a
      shallow tray at  450°C for  a minimum of 4  hr  to  remove  interfering
      organic substances.   Acidify by  slurrying 100 g  sodium  sulfate with
      enough ethyl  ether to just cover the  solid.  Add 0.1 ml concentrated
      sulfuric acid and  m-ix thoroughly.  ;Remove the ether under vacuum.
      Mix  1  g of the resulting solid with 5 ml  of  reagent water and
      measure the pH of  the mixture.   The pH must  be  below pH 4.  Store at
      130°C.

 7.4   SODIUM THIOSULFATE,  GRANULAR, ANHYDROUS — ACS  grade.

 7.5   SODIUM HYDROXIDE (NAOH), PELLETS — ACS grade.

      7.5.1   NaOH,  6 N —  Dissolve 216 g  NaOH in 900  mL  reagent water.

 7.6   SULFURIC ACID, CONCENTRATED — ACS  grade,  so. gr.  1.84.

      7.6.1   Sulfuric  acid,  12 N — Slowly  add  335 mL  concentrated
             sulfuric  acid to 665 mL of reagent water.

 7.7   POTASSIUM HYDROXIDE  (KOH),  PELLETS — ACS grade.

      7.7.1   KOH,  37%  (w/v)  — Dissolve 37  g KOH pellets in reagent water
             and  dilute  to 100 mL.

 7.8   CARBITOL (DIETHYLENE GLYCOL MONOETHYL ETHER) —  ACS grade.
      Available from Aldrich Chemical  Co.

 7.9   DIAZALD,  ACS  grade —  Available  from  Aldrich Chemical Co.

 7.10  DIAZALD SOLUTION —  Prepare a solution containing  10 g  Diazald in
      100 mL  of a  50:50  by volume mixture of ethyl ether and  carbitol.
      This solution  is stable for one  month  or  longer  when stored at 4°C
      in an  amber  bottle with a  Teflon-lined screw cap.

 7.11  TRIMETHYLSILYLDIAZOMETHANE  (TMSD) —  Available from Aldrich Chemical
      Co. as  a 2 molar solution  in hexane.   TMSD is stable during storage
      in this  solution.

 7.12  SODIUM  CHLORIDE  (NACL), CRYSTAL, ACS  GRADE — Heat treat in a
      shallow  tray  at 450°C  for  a minimum of 4  hr to remove interfering
      organic  substances.

 7.13  4,4'-DIBROMOOCTAFLUOROBIPHENYL (DBOB)  —  99% purity,  for use as
      internal  standard  (available from Aldrich  Chemical  Co).

 7.14  2,4-DICHLOROPHENYLACETIC ACID (DCAA)  — 99% purity, for use as
      surrogate standard (available from Aldrich Chemical Co).

7.15 MERCURIC  CHLORIDE -- ACS grade (Aldrich Chemical Co.)  - for use as a
      bacteriocide  (optional- see  Section 8).

7.16 REAGENT WATER — Reagent water is defined  as water that is
     reasonably free of contamination that  would prevent the

                              515.1-10

-------
      determination of any analyte of  interest.  Reagent water used to
      generate the validation data in  this method was distilled water
      obtained from the Magnetic Springs Water Co., Columbus, Ohio.

 7.17 SILICIC ACID, ACS GRADE.

 7.18 FLORISIL - 60-100/PR mesh (Sigma No. F-9127).  Activate by heating
      in a shallow container at 150°C  for at least 24 and not more than 48
      hr.

 7.19 STOCK STANDARD SOLUTIONS (1.00 /zg/pL) - Stock standard solutions
      may  be purchased as certified solutions or prepared from pure
      standard materials using the following procedure:

      7.19.1 Prepare stock standard solutions by accurately wei.ghing
             approximately 0.0100 g of pure material.   Dissolve the
             material  in MTBE and dilute to volume in  a  10-mL volumetric
             flask.   Larger volumes may be used at the convenience of the
             analyst.   If compound  purity is certified at 96% or greater,
             the  weight  may be used without correction to calculate the
             concentration of the stock standard.   Commercially prepared
             stock standards  may  be used at any concentration if they are
             certified by the manufacturer or by an  independent source.

      7.19.2  Transfer  the stock standard solutions  into  TFE-fluoro-
             carbon-sealed screw  cap  amber vials.   Store at  room  tempera-
             ture and  protect from  light.

      7.19.3  Stock standard solutions  should  be  replaced after  two months
             or. sooner if comparison with  laboratory fortified  blanks   or
             QC samples  indicate  a  problem.

7.20  INTERNAL STANDARD SOLUTION  - Prepare  an  internal  standard  solution  "
      by accurately weighing  approximately  0.0010 g of pure  DBOB
      Dissolve the DBOB in MTBE and dilute  to volume in  a 10-mL volumetric
      flask.  Transfer the  internal standard solution to  a TFE-fluoro-
      carbon-sealed screw  cap bottle and store at room temperature
      Addition of  25 /iL of the internal  standard solution to  10 mL of
      sample extract,  or  12.5 jjl to 5 mL of sample extract,  results in a
      final internal standard concentration of 0.25 /ig/mL.   Solution
      should be replaced when ongoing QC (Sect. 9) indicates  a problem.
      Note that DBOB has been shown to  be an effective internal standard
      for the method analytes, but other compounds may be used if the
     quality control requirements in Sect. 9 are met.

7.21 SURROGATE STANDARD SOLUTION - Prepare a surrogate standard solution
     by accurately weighing approximately 0.0010 g of pure DCAA
     Dissolve the DCAA in MTBE and dilute to volume in a 10-mL volumetric
     flask.   Transfer the surrogate standard solution  to a TFE-fluoro-
     carbon-sealed screw cap  bottle and store at room  temperature.
     Addition of  50 /zL of the surrogate standard solution to a 1-L sample
     prior to extraction results  in a surrogate standard concentration in
     the sample of 5 /ig/L and,  assuming quantitative recovery of DCAA, a
     surrogate standard  concentration in the final  extract  of 0.5 /ig/mL.

                              515.1-11

-------
          Solution should be replaced when ongoing QC (Sect.  9)  indicates a
          problem.  Note DCAA has been shown to be an effective  surrogate
          standard for the method analytes, but other compounds  may be used if
          the quality control requirements in Sect. 9 are met.

     7.22 LABORATORY PERFORMANCE CHECK SOLUTIONS — Prep.are a diluted dinoseb
          solution by adding 10 nl of the 1.0 M9/ML dinoseb stock solution to
          the MTBE and diluting to volume in a 10-mL volumetric  flask.  To
          prepare the check solution, add 40 til of the diluted dinoseb
          solution, 16 ML of the 4-nitrophenol stock solution,  6 /zL of the
          3,5-dichlorobenzoic acid stock solution, 50 p.1 of the  surrogate
          standard solution, 25 juL of the internal standard solution, and 250
          III of methanol to a 5-mL volumetric flask and dilute to volume with
          MTBE.  Methylate sample as described in Sects. 11.4 or 11.5.  Dilute
          the sample to 10 mL in MTBE.  Transfer to a TFE-fluorocarbon-sealed
          screw cap bottle and store at room temperature.  Solution should be
          replaced when ongoing QC (Sect. 9) indicates a problem.

8.   SAMPLE COLLECTION. PRESERVATION. AND STORAGE

     8.1  Grab samples must be collected in glass containers. - Conventional
          sampling practices (5) should be followed; however, the bottle must
          not be prerinsed with sample before collection.

     8.2  SAMPLE PRESERVATION AND STORAGE

          8.2.1  If residual chlorine is present,  add 80 mg of sodium
                 thiosulfate (or 50 mg sodium sulfite) per liter of sample to
                 the sample bottle prior to collecting the sample.

          8.2.2  After the sample is collected in  a  bottle containing the
                 dechlorinating agent, seal the, bottle and shake until
                 dissolved.

          8.2.3  The samples must be  iced  or refrigerated at 4°C away from
                 light from the time of collection  until extraction.  Pre-
                 servation study results  indicated  that most method analytes
                 present  in  samples were  stable for  14 days when stored  under
                 these conditions.   Analyte  stability may be affected by the
                 matrix;  therefore, the analyst should verify that the
                 preservation technique is  applicable to the samples  under
                 study.

          8.2.4  All performance data presented in  this method are from
                  samples  preserved with mercuric  chloride.  No suitable
                 preservation agent  (biocide)  has  been  found other than
                 mercuric-chloride.   However  the  use of mercuric chloride  is
                  not required due to  its  toxlcity  and potential  harm  to  the
                  environment.

          8.2.5   In  some  circumstances where  biological degradation  of target
                  pesticides  might be  expected,  use of mercuric chloride  may  be
                  appropriate to minimize  the  possibility of  false-negative
                  results.   If mercuric chloride  is to be used, add  it (Sect.

                                    515.1-12

-------
                  7.8)  to  the  sample  bottle  in  amounts to produce a
                  concentration of  10 mg/L.  Add  1 mL of a solution containing
                  10 mg/mL of  mercuric chloride in reagent water to the sample
                  bottle at the sampling site or  in the laboratory before
                  shipping to  the sampling site.  A major disadvantage of
                  mercuric chloride is that  it  is a highly toxic chemical;
                  mercuric chloride must be  handled with caution, and samples
                  containing mercuric chloride must be disposed of properly.

     8.3  EXTRACT STORAGE

          8.3.1   Extracts should be  stored  at 4°C away from light.
                  Preservation study  results indicate that most analytes are
                 stable for 28 days;  however, the analyst should verify
                 appropriate extract holding times applicable to the samples
                 under study.

9.  QUALITY CONTROL

     9_.l  Minimum quality control  (QC)  requirements are initial  demonstration
          of laboratory capability, determination of surrogate compound
          recoveries in each sample and blank,  monitoring internal  standard
          peak area or height in each sample and  blank (when internal  standard
          calibration  procedures are  being employed),  analysis of laboratory
          reagent blanks,  laboratory  fortified  samples,  laboratory  fortified
          blanks, and  QC samples.   A  MDL for each analyte must also  be
          determined.

     9.2  LABORATORY REAGENT BLANKS (LRB).   .Before processing any samples,  the
          analyst must demonstrate  that  all  glassware  and  reagent
          interferences are  under control.   Each  time  a  set  of samples  is
          extracted or reagents  are changed, a  LRB must  be analyzed.   If
          within  the retention time window of any analyte  the LRB produces  a
          peak that would  prevent the determination  of that  analyte, determine
          the  source of contamination and  eliminate  the  interference before
          processing samples.

     9.3  INITIAL DEMONSTRATION OF  CAPABILITY.

          9.3.1   Select a  representative fortified concentration  for each
                 analyte.  Suggested  concentrations  are  10 times  the EDL or a
                 concentration that represents  a  mid-point in  the calibration
                 range.  Prepare a  primary dilution  standard  (in methanol)
                 containing each analyte at  1000  times selected concentration.
                 With  a syringe, add  1 mL of the  concentrate  to each of four
                 to  seven  1-L  aliquots 'of reagent water, and  analyze each
                 aliquot according  to procedures  beginning in  Sect. 11.

          9.3.2   For each analyte the recovery value for all  of these samples
                 must  fall in  the range of ± 30%  of the fortified amount,  with
                 the RSD of the measurements 30% or less.  For those compounds
                 that meet the acceptable criteria, performance is considered
                 acceptable and sample analysis may begin.  For those
                 compounds that fail these criteria, this procedure must be

                                  515.1-13

-------
            reported using fresh samples' until satisfactory performance
            has been demonstrated.

     9.3.3  For each analyte, determine the MDL.  Prepare a minimum of 7
            LFBs at a low concentration.  Fortification concentrations in
            Table 3 may be used as a guide, or use calibration data
            obtained in Section 10 to estimate a concentration for each
            analyte that, will produce a peak with a 3-5 times signal to
            noise response.  Extract and analyze each replicate according
            to Sections 11 and 12.  It is recommended that these LFBs be
            prepared and analyzed over a period of several days, so that
            day to day variations are reflected in precision
            measurements.  Calculate mean recovery and standard deviation
            for each analyte. Use the standard deviation and the equation
            given in Table 3 to calculate the MDL.

     9.3.4  The initial demonstration of capability is used primarily to
            preclude a laboratory from analyzing unknown samples via a
            new, unfamiliar method prior to obtaining some experience
            with it.  It is expected that as laboratory personnel gain
            experience with this method the quality of data will improve
            beyond those required here.

9.4  The analyst is permitted to modify GC columns, GC conditions,
     concentration techniques (i.e., evaporation techniques), internal
     standard or surrogate compounds.  Each time such method
     modifications are made, the analyst must repeat the procedures in
     Sect. 9.3

9.5  ASSESSING SURROGATE RECOVERY.

     9.5.1  When surrogate recovery from a sample or method blank is <70%
            or >130%, check  (1) calculations to locate possible errors,
            (2) standard solutions for degradation, (3) contamination,
            and (4) instrument performance.  If those steps do not reveal
            the cause of the problem, reanalyze the extract.

     9.5.2  If a LRB extract reanalysis fails the 70-130% recovery
            criterion, the problem must [be identified and corrected
            before continuing.          '

     9.5.3  If sample extract reanaTysis meets the surrogate recovery
            criterion, report only data for the reanalyzed extract.  If
            sample extract continues to fail the recovery criterion,
            report all data for that sample as suspect.

9.6  ASSESSING THE INTERNAL STANDARD

     9.6.1  When using the internal standard calibration procedure, the
            analyst must monitor the IS response  (peak area or peak
            height) of all samples during each analysis day.  The IS
            response for any sample chromatogram should not deviate from
            the daily calibration check standard's IS response by more
            than 30%.                   j

                              515.1-14

-------
      9.6.2  If >30% deviation occurs with an individual  extract  optimize
             instrument performance and inject a second aliquot of that
             extract.

             9.6.2.1  If the reinjected aliquot produces  an acceptable
                      internal  standard response,  report  results for that
                      aliquot.

             9.6.2.2  If a  deviation  of greater than  30%  is obtained for
                      the reinjected  extract,  analysis  of the  samples
                      should be  repeated beginning  with Sect.  11,  provided
                      the sample is still  available.  Otherwise,  report
                      results  obtained  from  the reinjected  extract,  but
                      annotate as  suspect.

     9.6.3   If  consecutive  samples fail  the  IS response  acceptance
             criterion,  immediately analyze  a  calibration check  standard.

             9.6.3.1   If  the check standard  provides  a  response  within  20%
                      of  the predicted  value,,  then  follow procedures
                      itemized in Sect.  9.6.2  for each  sample  failing the
                      IS  response criterion.

             9.6.3.2   If  the check standard provides  a  response which
                      deviates more than 20% of  the predicted  value, then
                      the analyst must  recalibrate, as  specified in Sect.


9.7  ASSESSING LABORATORY PERFORMANCE  - LABORATORY FORTIFIED  BLANK

     9.7.1  The laboratory must analyze at least one laboratory fortified
            blank (LFB) sample with every 20 samples or one per sample
            set (all samples extracted within a 24-hr period)  whichever
            is greater.  The concentration of each analyte in  the LFB
            should be 10 times EDL or a concentration which represents a
            mid-point in the calibration.  Calculate accuracy  as percent
         •   recovery (X,-).   If the  recovery  of any  analyte  falls outside
            the control limits (see Sect. 9.7.2),  that  analyte is judged
            out of control,  and  the source of the  problem should be
            identified and  resolved before continuing analyses.

     9.7.2  Until  sufficient data .become available  from within their own
            laboratory, usually  a minimum of results  from 20 to 30
            analyses,  the laboratory  should  assess  laboratory  performance
            against  the control  limits  in Sect. 9.3.2 that  are derived
            from the data in Table, 2.   When  sufficient  internal
            performance data becomes  available,  develop control  limits
            from the mean percent  recovery (X)  and  standard deviation  (S)
            of the percent  recovery.  These  data are  used  to establish
           .upper  and  lower  control limits as  follows:

                    UPPER CONTROL  LIMIT  =  X  + 3S
                    LOWER CONTROL LIMIT  =  X  - 3S


                              515.1-15

-------
            After each five to ten new recovery measurements, new control
            limits should be calculated using only the most recent 20-30
            data points.  These calculated control limits should not
            exceed those established in Section 9.3.2.

     9.7.3  It is recommended that the laboratory periodically determine
            and document its detection limit capabilities for the
            analytes of interest.

     9.7.4  At least quarterly, analyze a QC sample from an outside
            source.

9.8  ASSESSING ANALYTE RECOVERY - LABORATORY .FORTIFIED SAMPLE MATRIX

     9.8.1  The laboratory must add a known concentration to a minimum of
            10% of the routine samples or one sample per set, whichever
            is greater.  The concentration should not be less then the
            background concentration of the sample selected for
            fortification.  Ideally, the concentration should be the same
            as that used for the laboratory fortified blank (Sect. 9.7).
            Over time, samples from -all routine sample sources should be
            fortified.

     .9.8.2  Calculate the percent recovery, P, of the concentration for
            each analyte, after correcting the analytical result, X, from
            the fortified sample for the background concentration, b,
            measured in the unfortified sample, i.e.,:

            P = 100 (X - b) / fortifying concentration,
                                       I
            and compare these values to control limits appropriate for
            reagent water data collected in the same fashion.  The value
            for P must fall between 6536-135% of'the fortified
            concentration.

     9.8.3  If the recovery of any such'analyte falls outside the
            designated range, and the laboratory performance for that
            analyte is shown to be in Control (Sect. 9.7), the recovery
            problem encountered with the fortified sample is judged to be
            matrix related, not system related.  The result for that
            analyte in the unfortified sample is labeled suspect/matrix
            to inform the data user that the results are suspect due to
            matrix effects.

9.9  ASSESSING INSTRUMENT SYSTEM - LABORATORY PERFORMANCE CHECK SAMPLE -
     Instrument performance should be monitored on a daily basis by
     analysis of the LPC sample.  The LPC sample contains compounds
     designed to monitor instrument sensitivity, column performance
     (primary column) and chromatographic performance.  LPC sample
     components and performance criteria are listed in Table 4.
     Inability to demonstrate acceptable instrument performance indicates
     the need for reevaluation of the instrument system.  The sensitivity
     requirements are set based on the EDLs published in this method.   If
     laboratory EDLs differ from those listed in this method,

                              515.1-16

-------
          concentrations of the LPC compounds must be adjusted to be
          compatible with the laboratory EDLs.

     9.10 The laboratory may adopt additional quality control practices for
          use with this method.  The specific practices that are most
          productive depend upon the needs of the laboratory and the nature of
          the' samples.  For example, field or laboratory duplicates may be
          analyzed to assess the precision of the environmental measurements
          or field reagent blanks may be used to assess contamination of
          samples under site conditions, transportation and storage.

10.   CALIBRATION AND STANDARDIZATION

     10.1 Establish GC operating parameters equivalent to those indicated in
          Sect.  6.10.  The GC system may be calibrated using either the
          internal standard technique (Sect.  10.2) or the external  standard
          technique (Sect. 10.3).  NOTE:  Calibration standard solutions must
          be prepared such .that no unresolved analytes are mixed together.

    .10.2 INTERNAL STANDARD CALIBRATION PROCEDURE — To use this approach, the
          analyst must select one or more internal standards compatible in
          analytical  behavior to the compounds of interest.  The analyst must
          further demonstrate that the measurement of the internal  standard is
          not affected by method or matrix interferences.  DBOB has been
          identified as a suitable internal standard.

          10.2.1 Prepare calibration standards at a minimum, of three
                 (recommend five) concentration levels for each analyte of
                 interest by adding volumes of one or more stock standards to
                 a volumetric,flask.  To each calibration standard,  add a
                 known constant amount of one or more of the internal
                 standards and 250 jiiL methanol, and dilute to volume with
                ,MTBE.  Esterify acids with diazomethane as described in Sect.
                 11.4 or 11.5.

                 Guidance on the number of standards is as follows:   A minimum
                 of three calibration standards are required to calibrate a
                 range of a factor of 20 in concentration.  For a factor of 50
                 use at least four standards, and for a factor of 100 at least
                 five standards.  One calibration standard should contain each
                 analyte of concern at a concentration 2 to 10 times greater
             .    than the method detection limit for that compound.   The other
                 calibration standards should contain each analyte  of concern
                 at concentrations that define the range of the sample analyte
                 concentrations or should define the working range  of the
                 detector.    .                     .

          10.2.2 Analyze each calibration standard according to the  procedure
                 (Sect.  11.9).   Tabulate response (peak height or area)
                 against concentration for each,compound and internal
                 standard.   Calculate the response factor (RF)  for  each
         .   , -'   analyte and surrogate using  Equation 1.    ,
                                   515.1-17

-------
             RF =
                     (As)
                     (Ais)  (Cs)

                 where:
                    Equation  1
                 c"
= Response for the analyte to be measured.
= Response for the internal  standard.
= Concentration of the internal  standard (/ig/L)
= Concentration of the analyte to be measured
   (M9/L).
     10.2.3 If the RF value over the working range is constant (20% RSD
            or less) the average RF can be used for calculations.  Alter-
            natively, the results can be used to plot a calibration curve
            of response ratios (As/Ajs) vs. Cs.

     10.2.4 The working calibration curve or RF must be verified on each
            working day by the measurement of a minimum of two calibra-
            tion check standards, one at the beginning and one at the end
            of the analysis day.  These check standards should be at two
            different concentration levels to verify the calibration
            curve.  For extended periods of analysis (greater than 8 hr),
            it is strongly recommended that check standards be inter-
            spersed with samples at regular intervals during the course
            of the analyses.  If the response for any analyte varies from
            the predicted response by more than + 20%, the test must be
            repeated using a fresh calibration standard.  If the results
            still do not agree, generate a new calibration curve.  For
            those analytes that failed the calibration verification,
            results from field samples analyzed since the last passing
            calibration should be considered suspect.  Reanalyze sample
            extracts for these analytes after acceptable calibration is
            restored.

10.3 EXTERNAL STANDARD CALIBRATION PROCEDURE

     10.3.1 Prepare calibration standard^ as described in Sect 10.2.1,
            omitting the use of an internal standard.

     10.3.2 Starting with the standard of lowest concentration, analyze
            each calibration standard according to Sect. 11.9 and tabu-
            late response (peak height or area) versus the concentration
            in the standard.  The results can be used to prepare a cali-
            bration curve for each compound.  Alternatively, if the ratio
            of response to concentration (calibration factor) is a con-
            stant over the working range (20% RSD or less), linearity
            through the origin can be assumed and the average ratio or
            calibration factor can be used in place of a calibration
            curve.

     10.3.3 The working calibration curve or calibration factor must be
            verified on each working day as described in Section 10.2.4.

                              515.1-18

-------
     10.4 Verify calibration standards periodically, recommend at least
          quarterly, by analyzing a standard prepared from reference material
          obtained from an independent source.  Results from these analyses
          must be within the limits used to routinely check calibration
11.   PROCEDURE
     11.1 MANUAL HYDROLYSIS, PREPARATION,  AND EXTRACTION.

          11.1.1 Add preservative(s)  (Sect.8) to LRBs and  LFBs.   Mark the
                 water meniscus on the side of the sample  bottle for later.
                 determination of sample volume (Sect.  11.1.9).   Pour the
                 entire sample into a 2-L  separatory funnel.   Fortify sample
                 with 50 fil of the surrogate standard solution.

          11.1.2 Add 250 g NaCl  to the sample,  seal,  and shake to dissolve
                 salt.

          11.1.3 Add 17 mL of 6  N NaOH to  the sample,  seal,  and  shake.   Check
                 the pH of the sample with  pH paper;  if the  sample does  not
                 have a pH greater than  or  equal  to  12, adjust the pH by
                 adding more 6 N NaOH.   Let  the sample  sit at  room temperature
                 for 1  hr,  shaking the separatory funnel and contents
                 periodically.   Note:  Since  many  of  the analytes  contained in
                 this method are applied as  a variety of esters  and  salts, it
                 is  vital  to hydrolyze them  to  the parent acid prior  to
                 extraction.   This step must  be  included in the  analysis  of
                 all  extracted  field  samples,  LRBs,  LFBs, LFMs,  and  QCS.

          11.1.4  Add  60 mL  methylene  chloride  to  the  sample bottle to rinse
                 the  bottle,  transfer  the methylene chloride to the  separatory
                 funnel  and  extract the sample  by vigorously shaking  the
                 funnel  for  2 min  with periodic venting to release excess
                 pressure.   Allow  the  organic layer to separate from  the water
                 phase  for  a minimum of 10 min.   If the emulsion  interface
                 between layers  is more than one-third the volume of  the
                 solvent layer,  the analyst must employ mechanical techniques
                 to complete the phase separation.  The optimum technique
                 depends upon the  sample, but may include stirring, filtration
                 through glass wool, centrifugation,  or other  physical
                methods.  Discard the methylene chloride phase (Sect. 14,15).

         11.1.5 Add a  second 60-mL volume of methylene chloride to the sample
                bottle and repeat the extraction procedure a second time,
                discarding the methylene chloride layer.   Perform a third
                extraction in the same manner.

         11.1.6 Add 17 mL of 12 N H2S04 to the  sample, seal, and  shake to
                mix.  Check the pH of the  sample with pH paper;  if the sample
                does not have a pH less than or equal  to 2,  adjust the pH by
                adding more 12 N H2S04.

         11.1.7 Add 120 mL ethyl ether to  the sample,  seal,  and  extract  the
                sample  by vigorously  shaking the funnel  for  2  min with

                                  515.1-19    -

-------
            periodic venting to release excess pressure.  Allow the
            organic layer to separate from the water phase for a minimum
            of 10 min.  If the emulsion interface between layers is more
            than one third the volume of the solvent layer,  the analyst
            must employ mechanical techniques to complete the phase
            separation.  The optimum technique depends upon  the sample,
            but may include stirring, filtration through glass wool,
            centrifugation, or other physical methods.  Remove the
            aqueous phase to a 2-L Erlenmeyer flask and collect the ethyl
            ether phase in a 500-mL round-bottom flask containing
            approximately 10 g of acidified anhydrous sodium sulfate.
            Periodically, vigorously shake the sample and drying agent.
            Allow the extract to remain in contact with the  sodium
            sulfate for approximately 2 hours.

     11.1.8 Return the aqueous phase to the separatory funnel, add a
            60-mL volume of ethyl ether to the sample, and repeat the
            extraction procedure a second time, combining the extracts in
            the 500-mL erlenmeyer flask:  Perform a third extraction with
            60 ml of ethyl ether in the!same manner.

     11.1.9 Determine the original.sample volume by refilling the sample
            bottle to the mark and transferring the water to a 1000-mL
            graduated cylinder.  Record'the sample volume to the nearest
            5 ml.

11.2 AUTOMATED HYDROLYSIS, PREPARATION, AND EXTRACTION. — Data presented
     in this method were generated using the automated extraction
     procedure with the mechanical separatory funnel shaker.

     11.2.1 Add preservative (Sect.  8.2) to any samples not  previously
            preserved, e.g., LRBs and LFBs.  Mark,the water  meniscus on
            the side of the sample bottle for later determination of
            sample volume (Sect. 11.2.9).  Fortify sample with 50 /zL of
            the surrogate standard solution.  If the mechanical
            separatory funnel shaker is used, pour the entire sample into
            a 2-L separatory funnel.  If the mechanical tumbler is used,
            pour the entire sample into a tumbler bottle.

     11.2.2 Add 250 g NaCl to the sample, seal, and shake to dissolve
            salt.

     11.2.3 Add 17 mL of 6 N NaOH to the sample, seal, and shake.  Check
            the pH of the sample with pH paper; if the sample does not
            have a pH greater than or equal to 12, adjust the pH by
            adding more 6 N NaOH.  Shake sample for 1 hr using the
            appropriate mechanical mixing device.  Note: Since many of
            the analytes contained in this method are applied as a
            variety of esters and salts, it is vital to hydrolyze them to
            the parent acid prior to extraction.  This step  must be
            included in the analysis of all extracted field  samples,
            LRBs, LFBs, LFMs, and QCS.   ;
                              515.1-20

-------
 11.2.4 Add 300 ml methylene chloride  to  the  sample  bottle  to  rinse
        the bottle,  transfer the methylene  chloride  to  the  separatory
        funnel  or tumbler bottle,  seal,  and shake  for  10  s,  venting
        periodically.   Repeat shaking  and venting  until pressure
        release is not  observed  during venting.  Reseal and  place
        sample  container in  appropriate mechanical mixing device.
        Shake or tumble the  sample for 1  hr.   Complete  and  thorough
        mixing  of the organic and  aqueous phases should be  observed
        at  least 2 min  after starting  the mixing device.

 11.2.5 Remove  the sample container from  the  mixing  device.  If the
        tumbler is used,  pour contents of tumbler  bottle  into  a 2-L
        separatory funnel.   Allow  the  organic layer  to  separate from
        the water phase for  a minimum  of  10 min.   If the  emulsion
        interface between layers  is more  than one  third the  volume of
        the solvent  layer, the analyst must employ mechanical
        techniques to complete the phase  separation.  The optimum
        technique depends upon the sample,  but may include  stirring,
        filtration through glass wool, centrifugation, or other
        physical  methods.  Drain and discard  the organic  p.hase.  If
        the tumbler  is  used,  return the aqueous phase to  the tumbler
        bottle.

 11.2.6  Add 17  ml of 12  N H2S04 to  the  sample, seal,  and shake  to
        mix.  Check the  pH of the  sample  with pH paper; if the sample
        does not  have a  pH less than or equal to 2,  adjust the pH by
        adding  more 12  N  H2S04.

 11.2.7  Add  300 mL ethyl  ether to  the  sample, seal,  and shake for 10
        s,  venting periodically.   Repeat  shaking and venting until
        pressure  release  is  not observed  during venting.  Reseal  and
        place sample container in  appropriate mechanical  mixing
        device.    Shake  or tumble sample for 1 hr.  Complete and
        thorough  mixing of the organic and  aqueous phases  should be
        observed  at least 2  min after  starting the mixing  device.

 11.2.8  Remove the sample container from  the mixing device.   If the
        tumbler is used,  pour  contents of tumbler bottle  into a 2-L
        separatory funnel.  Allow  the  organic layer to separate from
        the water  phase for  a minimum  of  10 min.   If the emulsion
        interface  between layers is more than one third the volume  of
        the solvent layer, the analyst must employ mechanical
        techniques to complete the phase  separation.   The  optimum
        technique depends upon the sample, but may include stirring,
        filtration through glass wool,  centrifugation,  or  other
       physical methods.  Drain and discard the aqueous phase.
       Collect  the extract  in a 500-mL round-bottom flask containing
       about 10 g of acidified anhydrous sodium sulfate.
       Periodically vigorously shake the sample and drying  agent.
       Allow the extract to remain in  contact with the sodium
       sulfate  for approximately 2 hr.

11.2.9 Determine the original sample volume by refilling  the sample
       bottle  to the mark and transferring  the water to a 1000-mL

                         515.1-21,

-------
            graduated  cylinder.   Record the  sample volume to the nearest
            5 ml.

11.3 EXTRACT CONCENTRATION

     11.3.1 Assemble a K-D concentrator by attaching a concentrator tube
            to  a 500-mL evaporative flask.

     11.3.2 Pour the dried extract through a funnel plugged with acid
            washed glass wool, and collect the extract in the K-D
            concentrator.  Use a  glass rod to crush any caked sodium
            sulfate during the transfer.  Rinse the round-bottom flask
            and funnel  with 20 to 30 ml of ethyl ether to complete the
            quantitative transfer.

     11.3.3 Add 1 to 2 clean boiling stones  to the evaporative flask and
            attach a macro Snyder column.  Prewet the Snyder column by
            adding about 1 mL ethyl ether to the top.  Place the K-D
            apparatus  on a hot water bath, 60 to 65°C, so that the
            concentrator tube is  partially immersed in the hot water, and
            the entire lower rounded surface of the flask is bathed with
            hot vapor.  At the proper rate of distillation the balls of
            the column will actively chatter but the chambers will not
            flood.  When the apparent volume of liquid reaches 1 mL,
            remove the K-D apparatus and allow it to drain and cool for
            at least 10 min.

     11.3.4 Remove the Snyder column and rinse the flask and its lower
            joint into the concentrator tube with 1 to 2 mL of ethyl
            ether.  Add 2 mL of MTBE-and a fresh boiling stone.  Attach a
            micro-Snyder column to the concentrator tube and prewet the
            column by  adding about 0.5 ml. of ethyl ether to the top.
            Place the  micro K-D apparatus on the water bath so that the
            concentrator tube is partially immersed in the hot water.
            Adjust the  vertical position of the apparatus and the water
            temperature as required to complete concentration in 5 to 10
            min.  When  the apparent volume of liquid reaches 0.5 mL,
            remove the  micro K-D from the bath and allow it to drain and
            cool.   Remove the micro Snyder column and add 250 p,i of
            methanol.   If the gaseous diazomethane procedure (Sect. 11.4)
            or trimethylsilyldiazomethane procedure (11.6) is used for
            esterification of pesticides; rinse the walls of the concen-
            trator tube while adjusting the volume to 5.0 mL with MTBE.
            If the pesticides will be esterified using the diazomethane
            solution (Sect. 11.5), rinse the walls of the concentrator
            tube while  adjusting the volume to 4.5 mL with MTBE.

11.4 ESTERIFICATION OF ACIDS USING GASEOUS DIAZOMETHANE — Results
     presented in  this method were generated using the gaseous diazomet-
     hane derivatization procedure.  See Section 11.5 and 11.6 for
     alternative procedures.

     11.4.1 Assemble the diazomethane generator (Figure 1) in a hood.
                              515.1-22

-------
     11.4.2  Add  5 mL  of  ethyl  ether  to  Tube  1.  Add  1 ml  of  ethyl ether,
            1  ml of carbitol,  1.5  mL of 37%  aqueous  KOH,  and 0.2 grams
            Diazald to Tube  2.   Immediately  place  the exit tube  into the
            concentrator tube  containing the sample  extract.   Apply
            nitrogen  flow (10  mL/min) to bubble diazomethane through the
            extract for  1 min.   Remove  first sample.  Rinse  the  tip of
            the  diazomethane generator  with  ethyl  ether  after methylation
            of each sample.  Bubble  diazomethane through  the second
            sample extract for 1 min.   Diazomethane  reaction mixture
            should be used to  esterify  only  two samples;  prepare new
            reaction  mixture in Tube 2  to esterify each  two  additional
            samples.   Samples  should turn yellow after  addition  of
            diazomethane and remain  yellow for  at  least  2 min.  Repeat
            methylation  procedure  if necessary.

     11.4.3  Seal concentrator  tubes  with stoppers.  Store at room
            temperature  in a hood  for 30 min.

     11.4.4  Destroy  any  unreacted  diazomethane  by  adding 0.1 to  0.2 grams
            silicic  acid to the concentrator tubes.   Allow  to stand until
            the  evolution of nitrogen gas has stopped (approximately  20
            min).  Adjust the  sample volume to  5.0. ml with  MTBE.

11.5 ESTERIFICATION  OF ACIDS USING DIAZOMETHANE SOLUTION —  Alternative
     procedure.
                            j*-
     11.5.1  Assemble  the diazomethane generator (Figure 2)  in a  hood.
            The  collection vessel  is a  10- or 15-mL  vial, equipped  with a
            Teflon-lined screw cap and  maintained  at 0-5C.

     11.5.2  Add  a sufficient amount of  ethyl ether to tube  1 to  cover the
            first impinger.  Add 5 mL of MTBE to  the collection  vial.
            Set  the  nitrogen flow at 5-10 mL/min.   Add  2 mL Diazald
            solution (Sect. 7.10)  and 1.5 mL of 37%  KOH solution to the
            second impinger.  Connect the tubing  as  shown and allow the
            nitrogen flow to purge the  diazomethane  from the reaction
            vessel  into  the collection  vial  for 30 min.  Cap the vial
            when collection is complete and  maintain at 0-5°C.  When
            stored at 0-5°C this diazomethane solution  may be used over a
            period of 48 hr.

     11.5.3 To  each concentrator tube containing  sample or  standard,  add
            0.5 mL diazomethane solution.   Samples should turn yellow
            after addition  of  the diazomethane solution and  remain yellow
            for at least 2  min.  Repeat methylation procedure if
            necessary.

     11.5.4 Seal concentrator  tubes with  stoppers.  Store at  room
            temperature  in  a  hood for  30  min.

     11.5.5 Destroy  any  unreacted diazomethane by adding 0.1  to 0.2 grams
            silicic  acid to the concentrator tubes.   Allow  to stand until
            the evolution of  nitrogen  gas has stopped (approximately 20
            min).  Adjust the  sample volume to 5.0 mL with  MTBE.

                               515.1-23

-------
 11.6  ESTERIFICATION  OF ACIDS  USING  TRIMETHYLSILYLDIAZOMETHANE  (TMSD)  --
      Alternative  procedure.   It  should  be  noted  that  the  gas
      chromatographic background  is  increased  when  TMSD  is  used  as  the
      derivatizing reagent  instead of the generated diazomethane.
      Although  no  method analyte  is  affected by this increased background,
      the  recommended surrogate,  2,4-dichloro-phenylacetic  acid,  is masked
      by an  interfering peak.   This  renders the surrogate  useless at  1
      //g/L or lower.  »Any compound found suitable when TMSD  is used is
      acceptable as a surrogate.

      11.6.1 Carry out  the  hydrolysis, clean-up,  and extraction  of  the
            method analytes as described up to Sect.  11.4.

      11.6.2 Add 50 fil  of the  2 M TMSD solution to  each  5 ml sample
            extract.

      11.6.3 Place the  tube containing the extract  into  a heating block at
            50°C  and heat the extract for 1 hour.

      11.6.4 Allow the  extract to cool to room temperature,- then add 100
            /zL of 2 M  acetic  acid in methanol  to react  any excess TMSD.

      11.6.5 Proceed with the  identification and measurement of the
            analytes using GC/ECD according to the procedures described
            in the method.

11.7 FLORISIL SEPARATION (optional)

     11.7.1 Place a small plug of glass-wool  into a 5-mL disposable glass
            pipet.  Tare the pipet,  and; measure 1 g of activated Florisil
          •  into the pipet.

     11.7.2 Apply 5 mL of 5 percent  metfianol  in MTBE to  the Florisil.
            Allow the liquid to just re^ch  the top of  the  Florisil.  In
            this and subsequent steps, allow the  liquid  level  to just
            reach  the top of the Florisfil  before  applying  the next  rinse,
            however,  do not allow the Florisil to go dry.   Discard
            eluate.

     11.7.3  Apply  5 mL methylated sample to the Florisil leaving silicic
            acid in the tube.   Collect eluate  in  K-D tube.

     11.7.4  Add 1  ml  of 5 percent methanol  in  MTBE  to  the  sample
            container,  rinsing walls.   Transfer the rinse  to the Florisil
            column leaving  silicic acid  in  the tube.   Collect  eluate  in a
            K-D tube.   Repeat  with 1-mL  and 3-mL  aliquots  of 5  percent
            methanol  in MTBE,  collecting eluates  in K-D  tube.

     11.7.5  If  necessary, dilute  eluate  to  10  mL  with  5  percent  methanol
            in  MTBE.

     11.7.6  Seal the  vial and  store  in  a refrigerator  if further process-
            ing  will  not be  performed  immediately.   Analyze  by GC-ECD.


                              515.1-24

-------
11.8 GAS CHROMATOGRAPHY

     11.8.1 Sect. 6.10 summarizes the recommended operating conditions
            for the GC.  Included in Table 1 are retention times observed
            using this method.  Other GC columns or chromatographic
            conditions may be used if the requirements of Sect.  9.3 are
            met.                               ..'-..".

     11.8.2 Calibrate qr verify the calibration of the system daily as
            described in Sect. 10.  The standards and extracts must be in
            MTBE.                                                      .

     11.8.3 If the internal  standard calibration procedure is used,
            fortify the extract with 25 . juL of internal standard solution.
            Thoroughly mix sample and place aliquot in a GC vial for
            subsequent analysis.

     11.8.4 Inject 2 nl of the sample extract.   Record the resulting peak
            size in area units.

    .11.8.5 If the response for the pe.ak exceeds the working range of the.
            system, dilute the extract and reanalyze.  If internal
            standard calibration  is used, add an additional amount of the
            IS, so that the amount in the diluted extract will match the
            calibration standards.                     .  .'

11.9 IDENTIFICATION OF ANALYTES

     11.9.1 Identify a sample component by comparison of its retention
            time to the retention time of a reference chromatogram.  If
            the retention time of an unknown compound corresponds, within
            limits, to the retention time of a  standard compound,  then
            identification is considered positive.

     11.9.2 The width of the retention time window used to make ,
            identifications should be based upon measurements of actual
            retention time variations of standards over the course of a
            day.  Three times the standard deviation .of a retention time
            can be used to calculate a suggested window size for a
            compound.  However, the experience  of the analyst should
            weigh heavily in the  interpretation of chromatograms.

     11.9.3 Identification requires expert judgement when sample
            components are not resolved chromatographically.  When GC
            peaks obviously represent more than one sample component
            (i.e., broadened peak with shoulder(s) or valley between two
            or more maxima,  or any time doubt exists over the
            identification of a peak on a chromatogram, appropriate
            alternative techniques, to help confirm peak identifica-
            tion, need to be employed.  For example, more positive
            identification may be made by the use of an alternative
            detector which operates on a chemical/physical principle
            different from that originally used, e.g., mass spectrom-


                              515.1-25

-------
                  etry,  or  the  use  of  a  second  chromatography  column.  A
                  suggested alternative  column  in  described  in Sect.  6.10.

12.  DATA ANALYSIS AND  CALCULATIONS

     12.1 Calculate  analyte  concentrations  in  the sample  from the  response  for
          the analyte using  the  calibration  procedure described  in Sect.  10.
          Use the multi-point  calibration to make all calculations.  Do not
          use the daily calibration verification  data to  quantitate  analytes
          in samples.

     12.2 If the  internal  standard calibration  procedure  is used,  calculate
          the concentration  (C)  in the  sample  using  the response factor (RF)
          determined in  Sect.  10.2 and  Equation 2, or determine  sample
          concentration  from the calibration curve.

                            (AS)(IS)
          C (/ig/L) = 	       Equation 2.
                           (Afs)(RF)(V0)    i

          where:

          As  = Response for the parameter'to be measured.
          Ais =  Response for the internal  standard.
          Is  = Amount of  internal  standard  added to each extract (/KJ).
          V0  - Volume of water extracted (L).
                                          j
     12.3 If the external  standard calibration  procedure  is used, calculate
          the amount of  material injected from  the peak response using the
          calibration curve  or calibration^factor determined  in Sect. JO.3.
          The concentration  (C)  in the  sample  can  be calculated from Equation
          3.

                         (A)(Vt)           ;
          C (fig/I) =	             Equation 3.
                         (V,)(V.)           ;

          where:

          A  = Amount of material  injected|(ng)
          V; =  Volume of extract injected
          Vt =  Volume of total  extract  (ill)
          Vs =  Volume of water extracted (mL).

13.  METHOD PERFORMANCE                   ;

     13.1 In a single laboratory, analyte recoveries from reagent water were
          used to determine  analyte MDLs, EDLs  (Table 3) and  demonstrate
          method range.   Analyte recoveries  and standard deviation about  the
          percent recoveries at one concentration  are given in Table 3. All
          data in Tables 1-3 were obtained using  diazomethane  for
          esterification.
                                   515.1-26

-------
      13.2  In a single laboratory, analyte recoveries from one standard
           synthetic ground waters were determined at one concentration level.
           Results were used to demonstrate applicability of the method to
           different ground water matrices.  Analyte recoveries from the one
           synthetic matrix are given in Table 2.

      13.3  The performance of dalapon using this method has been variable.
           Different users have had varying success in the accuracy and
           precision of dalapon measurements.  Because the dalapon methyl ester
           is much more volatile than the rest of the method analytes, it is
           suspected that extract concentration technique may be involved with
           poor recoveries of this analyte.   Therefore it is recommended that
           the analyst use caution to avoid losses due to volatization.

14.   POLLUTION PREVENTION

      14.1  This method uses significant volumes of organic solvents.  It is
           highly recommended that laboratories use solvent recovery systems to
           recover used solvent as sample extracts are being concentrated.
           Recovered solvents should be recycled or properly disposed of.

      14.2  For information about pollution prevention that may be applicable to
           laboratory operations, consult "Less is Better:  Laboratory Chemical
          Management for Waste Reduction" available from the American Chemical
           Society's Department of Government Relations and Science Policy,
           1155 16th Street N.W., Washington, D.C. 20036.

15.  WASTE MANAGEMENT

      15.1  It is the laboratory's responsibility to comply with all federal,
          state,  and local  regulations governing waste management, particu-
          larly the hazardous waste identification rules  and land disposal
          restrictions.   The laboratory using this method has the responsi-
          bility to protect the air,  water,  and land by minimizing and con-
          trolling all  releases from fume hoods and bench operations.   Compli-
          ance is also required with any sewage discharge permits and regula-
          tions.   For further information on waste management,  consult "The
          Waste Management Manual  for Laboratory Personnel," also available
          from the American Chemical  Society at the address in Sect.  14.2.

16.  REFERENCES

     1.    ASTM Annual  Book of Standards,  Part 11, Volume  11.02,  D3694-82,
          "Standard Practice for Preparation of Sample Containers and for
          Preservation,"  American  Society for Testing and Materials,  Philadel-
          phia,  PA,  p.  86,  1986.

     2.    "Carcinogens  -  Working with Carcinogens,"  Department of Health,
          Education,  and  Welfare,  Public Health Service,  Center for Disease
          Control,  National  Institute for Occupational  Safety and Health,
          Publication  No.  77-206,  Aug.  1977.
                                   515.1-27

-------
3.
4.
5.
"OSHA Safety and Health Standards, General  Industry,"  (29 CFR 1910),
Occupational Safety and Health Administration, OSHA 2206, (Revised,
January 1976).                      ,

"Safety in Academic Chemistry Laboratories," American  Chemical
Society Publication, Committee on Chemical Safety, "ivo. EoA\Ao\\,
1979.

ASTM Annual Book of Standards, Part 11, Volume 11.01,  D3370-82,
"Standard Practice for Sampling Water," American Society for Testing
and Materials, Philadelphia, PA, p. 130,  1986.
                              515.1-28

-------
    TABLES.  DIAGRAMS.  FLOWCHARTS.  AND VALIDATION DATA
               TABLE I.  RETENTION TIMES FOR METHOD ANALYTES
 Analvte
      Retention Time3
        (minutes)
Primary	  Confirmation
Dalapon
3,5-Dichlorobenzoic acid
4-Nitrophenol
DCAA (surrogate)
Dicamba
Dichlorprop
2,4-D
D80B (int. std.)
Pentachlorophenol (PCP)
Chloramben
2,4,5-TP
5-Hydroxydicamba
2,4,5-T
2,4-DB
Dinoseb
Bentazon
Picloram
DCPA acid metabolites
Acifluorfen
3.4
18.6
18.6
22.0
22.1
25.0
25.5
27.5
28.3
29.7
29.7
30.0
30.5
32.2
32.4
33.3
34.4 •
35.8
41.5
4.7
17.7
20.5
14.9
22.6
25.6
27.0
27.6
27.0
32.8
29.5
30.7
30.9
32.2
34.1
34.6
37.5
37.8
42.8
Columns and analytical conditions are described in Sect. 6.10.1
and 6.10.2.
                                 515.1-29

-------
t—
UJ
|
co
o
«=c

o:
UJ
H-
3c "a
r~H
z s-
i . i ni
UJ e f.
CO 4-> or
S to
LU 3
« 0
5™ .^
o -^
r< CO
UN -^
co c^
LU >>
H- oo
>•
— 1

z
•^
.v_^ O
g i. f^c
^J *• t/3
1 1 co
4-*
Z "*
0 3
CO •*->
O CO
eg «o
Q. CO
Q Q^
<
^i>
O
i u^
o c ^~
rj O D
i« 0 =3

»•*•
c£
0
5"
c5 co
c> ^^
GQ LU
cC H"
••J ^QC
LU 0
CO =3
ZO
CO CJ
CM CO
LU ">•
CQ 'to
*--C c*
i- <












ID
O
CSJ




CO
O
i— I







>^-
LO
i— H




i — 1
CSJ


CSJ
o








c:
co
s~
o
3
(J
< (












i — i— i LO i — cn
f^ C3 Lf) O C3 I-H
CO •— 1 CO CSJ




CSJ CSJ C3 CO <±> i— I
00 fH i— 1 CSJ CO
i-H i-H i — 1







CO ^" LO C3 i — 1 1 — *
10 »3- 1-^ o co cn
i— i i— i csj csj •— i




O i— l i— l O I — «3-
csj I-H co o co f-~



•— 1 O •— i O «^- O
i— <




00
CO
4-^
O
-0
to
4->
CO
c
co -o
c .0 -i-
o e e u
tsl to O CQ to .
fO S- Q Q.Q
»-> o i to i 
CT> CO C3
!__j







«* CO CO
CSJ IO O
CO »~H CSJ




LO CSJ 1 —
co o o


,.
O O csl




, -a
U
ro

U
camoa
5-Dichlorobenzoi
chlorprop
i~ « -r~* •
Ii CO Q (

i










«d- co co csj co CO co
co LO ^ cn co co 10




cn co r-~. «df- i — LO LO
co co csi co cn cn o
i-H r-H







CO LO IO CSJ LO «d- CO
•^ IO CO I-H LO IO O
r-H I— 1 CSI CO I-H I-H CO




CSI CO •— i O •— i r--. «*•
*j- o co co cn I-H co


S^^CSJ
O O i— i O O O O





a?
O
a.
^-^^
re |O
E CO
ro r— ^:
O O Q.
•r- C O
•o co s-
>>-f= 0
x Q.1— e Q.
QJ S- S-  ro O LO LO
0 >,-•- 4-> r— " -
1 — | | ICO •! — rt «
3 LO «:*• Q_ CX. CSI Csf





.
oo
CO

Q.
E
rO
oo
CO
1
1 —
t[
O
rO
CO
E
CO

^_)

4_9
c
O)
00
CO
i-
Q-
co


•o
to
^/
to
.JE

CO
0
0)
CO
•o

4^
c
3
0
re

o
OJ_
Data corrected

<0







































S_
CO
>
0
o
CO
4-5
e
CO
0
s_
CO
Q.
CO
en
10
co
ro
n
CtL

XI
































^
O)
o
(J
CO
s-
co

O)
o.
CO
-C
•*-*
f 1
Q

_
o
1e
*rr
m
SR = standard di

U

-i= '
E
S-
**"
CO ,

^
ro

O

S-
co
ro
-^
C
"^
CL
OO
g—
to

00
co

s_

i— =3
_iO O
E
cZ
•a
£Z C
3 -I—
o
^ >>
C Q.
3 £
O O
E O
Corrected for a
Absopure Water

•o
515.1-30

-------
                                                                                                                                      1
 s
         u

         Q
O UJ
   CJ3


O LU
 O
 CU
ce:
co I—
LU
a:
>- o
O —I
o a
«C LU

>•
a; co
o i—
a: HH
o _i
CO
* i— 1 i — 1
o o r-. «=f
en en co r*^
1 — . to to to

0 O
CVJ o O cvj
O •— i «*• O

00
CU
.4J
^z
o
ro
+J
0)
- E

^3
•i»
c o
O CO ro
Q Q. Q
1 cO 1 

to
0
0
U
ro

0
o
N
ai
J3
0
i.
o
r—

U

O
LO
^
CO

o
«5f
o

cvj
CVJ
o
lO
1— 1
2
-

o
«<•
o









a.
0
s-
o.
$-
o

o

Q

o
*3"
0

CO
CO
o

CO
cvj
to

o
*3-
o













f™i
QJ
00
o

•1—
o
0

o
o

i^H
C3
o
CO
CVJ
s
1 —
o

0
o





ro
E
ro
(J

T3

X
o

•a_
~T"
1
LO

O
CVJ
O

^^
r-H
O
LO
i-H
2
r-.

0
cvj
O







r—
O
rz
(U

CL
O

4->
^^
1


O
r-H
O
CVJ
CO
0
0


3
to

0
1— 1
0
a.
o
ex.
*-*
, 	
o
CU

a.
o

o
r—
.C
O
ro
+J

O
D_

LO
i—)
O

LO
I— t
0


2
-

CVJ
1— 1
o












E
CO
1-
O
13

a.

0 — (
CO CVJ
o o
LO
^ i-H
O CVJ
o o
i-H CO
cvj cvj
sS
I-*- vp
o
CO O
O CVJ
0 0












Q.
r— |—
1 |
LO LO
^~ ^"
•* - •»
cvj cvj
                                                                                       (U
                                                                                       (U

                                                                                       en
                                                                                       cu
                                                                                                                 •-I  CU £=
                                                                                                              s- a; en
                                                                                                              ro    •>-
                                                                                                              a.  i  oo

                                                                                                              0£. +J JZ

                                                                                                              o's-t^
r—
0)
0)
>~~
CU
Sr
c.
0)
•o
iH
c
o
u
c u
O O re
•t-> -r- i.
CQ O X
a> 
•i- CO r—
•O Q ro
CU "O -r-
Q. O M-

                                                                                                              _

                                                                                                              Q  0) C
                                                                                                              ~
                                                                                                              i.
                                                                                                              a>
                                                                                                                       a>
















^

^J c?
frt *
O
(J
•r— H
i — a>
a. -g^
S- "S
<+- *~.
O "-
1
s- £ "
jU +J OJ
E co «
= II ^
-^^_ II
i- ^^
o c
^"" *o

2J cu
= •»->
• ro
re 0
•r—
•^ Q.
tn 
c> S- CU
O *XU
o ^^
O ^D
co i- ro
•e cu -a
5- -Q c
ca E ro
1 3 4->
. c: oo
^ n n

4-> c: co




•r- C -C
CU O re
00 4-> CD-I—
ro ro c
C -^- CU
•0 ••- -0 3
0) E i— r—
C t CU ro
•- 0) -r- >
0) 0) S-
•a a CD a>
i — >
• «  .c E -C
•r- +j re o
E co -i—
c c •>
o cu -i- LO
•r- S-
4-> 3 -0 >>
O T3 C r—
CU CU =3 CU
+-> (J O 4->
CUOQ.ro
-a i- E E
Q. 0 -f-
•CJ OX
cu -a o
+-> c t- s-
rO re O CO.
E Q-
•r- c: i— ro
+-> o cu
oo -i- > 4-
CU -(-> CU O
II C "~ 0
•r- rO -i—
— J M— +->
Q 0) S- re
LU Q 0 i-
                                                 515.1-31

-------


















o
^4
.3
o
00
I
UJ
C_>
1
Ul
Q.
>-
cn
1
o
CO
 s- -,=• £
H JS -^ Q +J
^••1 O *p"» i ••_
1 = z 4?
> CO l<_
4J S_ S_
••- 01 at
> o a.
•M CO c
•r- E E
£ 2 =
cu .c o
1
c
0
•r™
-M
CO
^5
o-
cu
cu
1—
5
<=
01
cu
CO
3
u

CO'
c_>


sJ
o ^
CO •— ' -^
'"-3 2
«3» ^^
CO X ^H
•l~ 5**
01 00 ;s
01 00
co T~>
43 II
~^ , ,
CO ~~
a. a-
II
u.
!o-
• ••">_><_> ,j









*
+J
en
•f—
o>
^:
J=
+->
c
cu
-M ,

4->
CO
^=
+J
T3
'5
CO
cu
Q.
cu
4^
01

'"~ c
o
1 5
p^r s.
C- °^
5- a,
••" ^
* ^
-M ^
f |
^ O)
IT?
** *


5 co
±: cu
5 °L
3 o
2
« ^
2. *
••a) +*
•C ^
•^ 5
-' 1
^-s *
CM -°
^^** £Z
d. °
3 •£ ^,-
* r^
| %
^ cS c
A

JZ
4-1
•o
'5
CO
cu
CL
CU
01
CO

CD
CO
cu
r~
4->
01
If—
3:

^-•j
c
CO
01
CO
CU
a.
§
4->
^r
4->

c
cu
cu
-M
O)
*
(J
cu
£= ,
CU CO
S cu
'•i- a_
o
1*
4-> 
• CO
01
•i— *\
__ -M 01
S o-
cu cu
5»- O1
i cu
* -i-^

515.1-32

-------
Nitrogen
    Tub«1
                 Tub«2
                              Samplt
                              Tub*
  FIGURE 1.  GASEOUS DIAZOJ1ETHANE GENERATOR
                       515.1-33'

-------

\
           ^

      \\
I
                 5*. •.»
                               i.
                               4»



                              J
                              O U
                               o
                   I?
                        (VI
                       i){
                         4J
                                      UJ
                                      O
                                      UJ


                                      s

                                      I—i
                                      o

                                       •
                                      CM
               T
                  • •
            515.1-34

-------
           METHOD 515.2.
DETERMINATION OF CHLORINATED ACIDS IN WATER
USING LIQUID-SOLID EXTRACTION AND GAS
CHROMATOGRAPHY WITH AN ELECTRON CAPTURE DETECTOR
                                 Revision 1.1


                          Edited by J.W. Munch (1995)




R.C. Dressman and J.J. Lichtenberg - EPA 600/4-81-053, Revision  1.0  (1981)

J.W. Hodgeson - Method 515, Revision 2.0 (1986)

T. Engel (Battelle Columbus Laboratories) - National Pesticide Survey
Method 3, Revision 3.0 (1987)

R.L. Graves - Method 515.1, Revision 4.0 (1989)

J.W. Hodgeson - Method 515.2, Revision 1.0 (1992)
                    NATIONAL EXPOSURE RESEARCH LABORATORY
                      OFFICE OF RESEARCH AND DEVELOPMENT
                     U.S.  ENVIRONMENTAL  PROTECTION AGENCY
                           CINCINNATI, OHIO  45268
                                   515.2-1

-------
                                METHOD 515.2

              DETERMINATION OF CHLORINATED ACIDS IN WATER USING
                LIQUID-SOLID EXTRACTION AND GAS CHROMATOGRAPHY
                       WITH AN ELECTRON CAPTURE DETECTOR
1.    SCOPE AND APPLICATION
     1.1
     1.2
     1.3
     1.4
This is a gas chromatographic (GC) method applicable to the determi-
nation of certain chlorinated acids in ground water and finished
drinking water.  The following compounds can be determined by this
method:
                                            Chemical  Abstract Services
                                                  Registry Number

                                                    50594-66-6
                                                    25057-89-0
                                                       94-75-7
                                                       94-82-6

                                                     1918-00-9
                                                    '   51-36-5  '
                                                      120-36-5
                                                       88-85-7
                                                     7600-50-2
                                                       87-86-5
                                                     1918-02-1
                                                       93-76-5
                                                       93-72-1
Analvte

Acifluorfen
Bentazon
2,4-D
2,4-DB
Dacthal acid metabolites l
Dicamba
3,5-Dichlorobenzoic acid
Dichlorprop
Dinoseb
5-Hydroxydicamba
Pentachlorophenol (PCP)
Picloram
2,4,5-T
2,4,5-TP(Silvex)

(a>  Dacthal monoacid and diacid metabolites included in method
scope; Dacthal diacid metabolite used for validation studies.

This method is applicable to the[determination of salts and esters
of analyte acids.  The form of each acid is not distinguished by
this method.  Results are calculated and reported for each listed
analyte as the total free acid.

Single.laboratory accuracy and precision data and method detection
"limits (MDLs) have been determined for the analytes above (Sect.
13).  Observed detection limits may vary among water matrices,
depending upon the nature of interferences in the sample matrix and
the specific instrumentation used.  ,                     .     .

This method is restricted to use by or under the supervision of
analysts experienced in the use of GC and in the interpretation of
gas chromatograms.  Each analyst must demonstrate the ability to
generate acceptable results with this method using the procedure
described in Sect. 9.3.
                                    515.2-2

-------
     1.5  Analytes that are not separated chromatographically, (i.e., have
          very similar retention times) cannot be individually identified and
          measured in the same calibration mixture or water sample unless an
          alternative technique for identification and quantitation exists
          (Sect. 11.7).

     1.6  When this method is used to analyze unfami!iar samples for any or
          all of the analytes above, analyte identifications should be con-
          firmed by analysis on a second gas chromatographic column or by gas
          chromatography/mass spectrometry (GC/MS).

2.   SUMMARY OF METHOD

     2.1  A 250-mL measured volume of sample is adjusted to pH 12 with 6 N
          sodium hydroxide for 1 hr to hydrolyze derivatives. ( Note: Since
          many of the analytes contained in this method are applied as a
          variety of esters and salts, it is,vital to hydrolyze them to the
          parent acid prior to extraction.)  Extraneous organic material is
          removed by a-solvent wash.  The sample is acidified, and the chlori-
          nated acids are extracted with a 47 mm resin based extraction disk.
          The acids are converted to their methyl esters using diazomethane or
          alternatively, trimethylsilyldiazomethane (TMSD).  Excess derivatiz-
          ing reagent is removed, and the esters are determined by capillary
          column GC using an electron capture detector (ECD).  Analytes are
          quantitated using procedural standard calibration (Sect. 3.14).

3.   DEFINITIONS

     3.1  INTERNAL STANDARD (IS) -- A pure analyte(s) added to a sample,
          extract,  or standard solution in known amount(s), and used to
          measure the relative responses of other method analytes and surro-
          gates that are components of the same sample or solution.  The IS
          must be an analyte that is not a sample component.

     3.2  SURROGATE ANALYTE (SA) — A pure analyte(s),  which is extremely
          unlikely to be found in any sample, and which is added to a sample
          aliquot in known amount(s) before extraction or other processing,
          and is measured with the same procedures used to measure other
          sample components.   The purpose of the SA is to monitor method
          performance with each sample.

     3.3  LABORATORY DUPLICATES (LD1 AND LD2) — Two aliquots of the same
          sample taken in the analytical  laboratory and analyzed separately
          with identical  procedures.  Analyses of LD1 and LD2 indicate the
          precision associated with laboratory procedures,  but not with sample
          collection,  preservation,  or storage procedures.

     3.4  FIELD DUPLICATES (FD1 AND FD2)  — Two separate samples collected at
          the same  time  and place under identical  circumstances and treated
          exactly the same throughout field and laboratory procedures.
          Analyses  of FD1  and FD2 give a measure of the precision associated


                                    515.2-3

-------
     with sample collection, preservation and storage, as well as with
     laboratory procedures.

3.5  LABORATORY REAGENT BLANK  (LRB) — An aliquot of reagent water or
     other blank matrix that is treated exactly as a sample including
     exposure to all glassware, equipment, solvents, reagents, internal
     standards, and surrogates that are used with other samples.  The LRB
     is used to determine if method analytes or other interferences are
     present in the laboratory environment, the reagents, or the appara-
     tus.

3.6  FIELD REAGENT BLANK (FRB) — An aliquot of reagent water or other
     blank matrix that is placed in a sample container in the laboratory
     and treated as a sample in all respects, including shipment to the
     sampling site, exposure to sampling site conditions, storage,
     preservation and all analytical procedures.  The purpose of the FRB
     is to 'determine if method analytes or other interferences are
     present in the field environment.

3.7  INSTRUMENT PERFORMANCE CHECK SOLUTION (IPC) — A solution of one or
     more method analytes, surrogates, internal standards, or other test
     substances used to evaluate the performance of the instrument system
     with respect to a defined set of criteria.

3.8  LABORATORY FORTIFIED BLANK (LFB) — An aliquot of reagent water or
     other blank matrix to which known quantities of the method analytes
     are added in the laboratory.  The LFB is analyzed exactly like a
     sample, and its purpose is to determine whether the methodology is
     in control, and whether the laboratory is capable of making accurate
     and precise measurements.

3.9  LABORATORY FORTIFIED SAMPLE MATRIX (LFM) — An aliquot of an envi-
     ronmental sample to which known quantities of the method analytes
     are added in the laboratory.  The LFM is analyzed exactly like a
     sample, and its purpose is to determine whether the sample matrix
     contributes bias to the analytical results.  The background concen-
     trations of the analytes  in the sample matrix must be determined in
     a separate aliquot, and the measured values in the LFM corrected for
     background concentrations.         >

3.10 STOCK STANDARD SOLUTION (SSS) — A concentrated solution containing
     one or more method analytes prepared in the laboratory using
     assayed reference materials or purchased from a reputable commercial
     source.

3.11 PRIMARY DILUTION STANDARD SOLUTION (PDS) — A solution of several
     analytes prepared in the  laboratory from stock standard solutions,
     and diluted as needed to  prepare calibration solutions and other
     needed analyte solutions.

3.12 CALIBRATION STANDARD (CAL) — A solution prepared from the primary
     dilution standard solution or stock standard solutions and the

                               515.2-4

-------
           internal  standards and surrogate analytes.   The CAL solutions are
           used to calibrate the instrument response with respect to analvte
           concentration.        •                                     anaiyue

      3.13  QUALITY CONTROL SAMPLE (QCS)  -  A solution  of method analytes of
           known  concentrations  which  is used to  fortify an  aliquot  of LRB  or
           sample matrix.   The QCS  is  obtained from  a  source external  to the
           laboratory  and  different  from the source  of calibration standards.
           it  is  used  to check laboratory performance  with externally  prepared
           test materials.

      3.14  PROCEDURAL  STANDARD CALIBRATION  -  A  calibration  method  where
           aqueous calibration standards  are  prepared  and  processed  (e  q
           purged  extracted,  and/or  derivatized)  in  exactly  the  same manner  as
           a sample.   All  steps  in the process  from  addition  of  samplinq
           preservatives through  instrumental  analyses  are included  in  the
           calibration.  Using procedural standard calibration compensates for
           any  inefficiencies in  the processing procedure.

4.    INTERFERENCES

     4.1  Method interferences may be caused by contaminants in solvents
          reagents, glassware and other sample processing apparatus that'lead
          to discrete artifacts  or elevated baselines  in gas chromatograms
          All  reagents and apparatus must be routinely demonstrated to be free
          from interferences under analytical conditions by analyzing labora-
          tory reagent blanks as described  in Sect.  9.2.

          4.1.1  Glassware must  be scrupulously cleaned. (1)  Clean  all glass-
                 ware as  soon as possible after use  by thoroughly rinsing with
                 the last  solvent used in it.  Follow by washing with  hot
       ;          water and detergent and thorough rinsing with dilute  acid
                 tap and  reagent water.   Drain dry,  and heat in an  oven or'
                 muffle furnace  at  400°C for 1 hr.   Do not heat volumetric
                 glassware.   Thermally stable materials such as PCBs might  not
                 be  eliminated by this treatment.  Thorough  rinsing  with
                 acetone may be  substituted for the  heating.   After  glassware
                 is  dry and  cool,  store  it  in a clean  environment to prevent
                 any accumulation of dust or other contaminants.  Store in-
                 verted or capped with  aluminum foil.

         4.1.2  The use of  high  purity  reagents  and solvents  helps  to  mini-
                 mize  interference problems.   Purification of solvents  by
                 distillation in  all-glass  systems may  be  required.
                 WARNING:  When  a solvent is  purified,  stabilizers and  preser-
                 vatives added by the manufacturer are  removed,  thus poten-   •
                 tially making the solvent  hazardous and reducing the shelf
                 1 i f e.

    4.2  The acid forms of the analytes are  strong  organic acids which react
         readily with alkaline substances  and can be lost during sample
         preparation.  Glassware and glass  wool must be acid-rinsed with 1 N

                                   515.2-5

-------
     hydrochloric acid and the sodium sulfate must be acidified with
     sulfuric acid prior to use to avoid analyte losses due to adsorp-
     tion.

4.3  Organic acids and phenols, especially chlorinated compounds,  cause
     the most direct interference with the determination.   Alkaline
     hydrolysis and subsequent extraction of the basic sample removes
     many chlorinated hydrocarbons and phthalate esters that might
     otherwise interfere with the electron capture analysis.

4.4  Interferences by phthalate esters can pose a major problem in pesti-
     cide analysis when using the ECD.  Phthalates generally appear in
     the chromatogram as large peaks.  Common flexible plastics contain
     varying amounts of phthalates,'that are easily extracted or leached
     during laboratory operations.  Cross-contamination of clean glass-
     ware routinely occurs when plastics are handled during extraction
     steps, especially when solvent-wetted surfaces are handled.  Inter-
     ferences from phthalates can best be minimized by avoiding the use
     of plastics  in the laboratory.   Exhaustive purification of reagents
     and glassware may be required to eliminate background phthalate
     contamination. (2,3)

4.5  Interfering  contamination may occur when a sample containing low
     concentrations of analytes is analyzed  immediately following a
     sample containing relatively high concentrations  of analytes.
     Between-sample rinsing of the sample syringe  and  associated equip-
     ment with methyl-tert-butyl-ether  (MTBE) can  minimize  sample cross-
     contamination.  After  analysis  of a  sample containing  high concen-
     trations of  analytes,  one or more  injections  of  MTBE  should be made
     to ensure that accurate  values  are  obtained  for  the next  sample.

4  6  Matrix  interferences may be  caused  by contaminants that  are coex-
     tracted  from the  sample.  Also,  note that  all  analytes  listed  in  the
     Scope  and Application  Section are  not resolved from each  other  on
     any  one  column,  i.e.,  one analyte  of interest may interfere with
     another  analyte  of  interest.  The  extent of  matrix interferences
     will  vary considerably from  source  to source,  depending  upon  the
     water  sampled.   The  procedures  in  Sect.  11  can be used to overcome
     many of  these interferences.  Analyte  identifications should  be
     confirmed  (Sect.  11.7).

4  7  Gas  chromatographic  background  is  significantly increased when TMSD
      is  used  as  the  derivatizing  reagent instead  of the generated  diazo-
     methane.  Although  no  method analyte is affected by this increased
      background,  the  recommended  surrogate,  2,4-dichloro-phenylacetic
      acid,  is masked  by  an  interfering  peak.  This renders the surrogate
      useless  at  1 [ig/L or lower.   Any compound  found suitable when TMSD
      is  used is  acceptable as a  surrogate.

4.8   It  is important that samples and working standards be contained in
      the same solvent'.  The solvent  for working standards  must be the
      same as the final solvent used  in  sample preparation.  If this is

                               515.2-6

-------
           not the case, chromatographic comparability of standards'to sample
           extracts may be affected.

 5.    SAFETY

      5.1   The toxicity or carcinogenicity of each reagent used in this method
           has not been precisely defined; however,  each chemical  compound must
           be treated as a potential health hazard.   Accordingly,  exposure to
           these chemicals must be reduced to the lowest possible  level.   The
           laboratory is responsible for maintaining a current  awareness  file
           of OSHA regulations regarding the safe handling of the  chemicals
           specified in this  method.  A reference file of material  safety data
           sheets should also be made available  to all  personnel  involved in
           the chemical  analysis.   Additional  references to laboratory safety
           are available and  have  been identified (5-7)  for the information of
           the analyst.

      5.2   DIAZOMETHANE — A  toxic carcinogen  which  can  explode under  certain
           conditions.   The following precautions must  be followed:

           5.2.1   Use the  diazomethane generator behind  a safety shield in  a
                  well  ventilated  fume hood.   Under  no  circumstances can  the
                  generator be heated above 90°C,  and  all 'grinding  surfaces
                  such  as  ground glass joints, sleeve  bearings, and glass
                  stirrers must  be avoided.  Diazomethane solutions must  not  be
                  stored.   Only  generate  enough  for  the  immediate needs.  The
                  diazomethane generator  apparatus used  in  the  esterification
                  procedure (Sect.  11.4)  produces  micromolar  amounts of diazo-
                  methane  in  solution  to  minimize  safety  hazards.   If the
                  procedure is  followed exactly,  no  possibility for explosion
                  exists.

      5.3  METHYL-TERT-BUTYL  ETHER -  Nanograde,  redistilled  in glass,  if
          necessary.  Must be  free  of peroxides  as  indicated by EM Quant test
          strips  (available  from  Scientific Products Co.,  Cat.  No. PI 126-8,
          and other  suppliers).

      5.4  WARNING:   When  a solvent  is  purified,  stabilizers  added by the
          manufacturer are removed, thus  potentially making  the solvent
          hazardous.

6.   EQUIPMENT AND SUPPLIES  (All  specifications are suggested.   Catalog
     numbers  are included for  illustration only.)

     6.1  KONTES FILTER FUNNELS —  Fisher Cat. No. 953755-0000  or equivalent.

     6.2  VACUUM FLASKS — 1000 mL with glass side arm

     6.3  VACUUM MANIFOLD —  The manifold should be capable of  holding 6-8
          filter flasks in series with house vacuum.  Commercial  manifolds are
          available from a number of suppliers,  e.g.,  Baker, Fisher,  and
          Vari an.

                                    515.2-7

-------
6.4  CULTURE TUBES (25 x 200 mm) WITH TEFLON-LINED SCREW CAPS — Fisher
     Cat. No. 14-933-1C, or equivalent.

6.5  PASTEUR PIPETS — Glass disposable (5 mL)

6.6  LARGE VOLUME PIPETS — Disposable, Fisher Cat. No. 13-678-8 or
     equivalent.

6.7  BALANCE — Analytical, capable of weighing to .0001 g.

6.8  pH METER — Wide range capable of accurate measurements in the pH =
     1-12 range.

6.9  DIAZOMETHANE GENERATOR — See Figure 1 for a diagram of an all glass
     system custom made for these validation studies.  A micromolar
     generator is also available from Aldrich Chemical.

6.10 ANALYTICAL CONCENTRATOR — Six or twelve positions, Organomation N-
     EVAP Model No. 111-6917 or equivalent.

6.11 GAS CHROMATOGRAPHY — Analytical system complete with gas chromato-
     graph equipped with ECD, split/splitless capillary injector, temper-
     ature programming, differential flow control and  all required acces-
     sories.  A data system is recommended for measuring peak areas.  An
     autoinjector is recommended to improve precision  of analysis.

6.12 GC COLUMNS AND RECOMMENDED OPERATING CONDITIONS

     6.12.1 Primary — DB-5 or equivalent, 30 m x  .32  mm ID, 0.25 /im film
            thickness.  Injector Temp. = 200°C, Detector Temp. - 280°C,
            Helium linear Velocity is 30 cm/sec at 200°C and 10 psi, 2 /*L
            splitless injection with purge on 3 min.   Program:  Hold at
            60°C 1 min.;, increase to 260°C at 5°C/min. and hold 5 min.

     6.12.2 Confirmation — DB-1701 or equivalent, 30  m x  .32 mm ID, 0.25
            im film thickness.  Injector Temp. = 200  °C, Detector Temp. =
            280°C, Helium linear velocity is 30 cm/sec at 200°C and 10
            psi, 2 ML splitless injection with purge  on 3 min. Program:
            Hold at 60°C 1 min., increase to 260°C at  5°C/min. and hold 5
            min.                        '•

6.13 GLASS WOOL — Acid washed with IN Htl and heated  at 450°C for 4 hr.

6.14 SHORT RANGE pH PAPER  (pH=0-3).

6.15 VOLUMETRIC FLASKS -- 50 mL, 100 mL, and 250 mL

6.16 MICROSYRINGES — 25 fil, 50 p.L, 100 /tL, 250 ML,  500 /d-   '

6.17 AMBER BOTTLES — 15 mL, with Teflonplined screw  caps

6.18 GRADUATED CYLINDER — 250 mL

                               515.2-8

-------
     6.19 SEPARATORY FUNNEL — 500 ml

     6.20 GRADUATED CENTRIFUGE TUBES - 15 mL or 10 mL Kuderna Danish Concen-
          trator tubes

7.   REAGENTS AND STANDARDS

     7.1  EXTRACTION DISKS, 47 mm — Resin based polystyrenedivinylbenzene

     7.2  REAGENT WATER — Reagent water is defined as a water in which an
          interference is not observed at the MDL of each analyte of interest.

          7.2.1  A Mi 11ipore Super-Q water system or its equivalent may be
                 used  to  generate deionized reagent water.   Distilled water
                 that  has been passed through granular charcoal  may also be
                 suitable.

          7.2.2  Test  reagent water each day it  is  used by  analyzing according
                 to Sect.  11.

     7.3   METHANOL  —  Pesticide quality  or equivalent.

     7.4   METHYL-TERT-BUTYL ETHER (MTBE)  -  Nanograde,  redistilled  in  glass if
          necessary.   Ether must  be demonstrated  to  be  free  of peroxides   One
          test  kit  (EM Quant Test Strips),  is  available  from EM  Science/
          Gibbstown, NJ.   Procedures  for removing peroxides  from the  ether are
          provided  with the test  strips.   Ethers must be  periodically  tested
          (at least monthly)  for  peroxide  formation  during use.   Any  reliable
          test  kit  may be used.

    7.5   SODIUM SULFATE -  (ACS)  GRANULAR, ANHYDROUS -  Heat in a shallow
          tray  at 400 C for a minimum of 4 hr to remove phthalates and other
          interfering organic  substances.  Alternatively, extract with methy-
          lene  chloride in  a Soxhlet  apparatus for 48 hr.  After  cleaning
          store in a glass  (not plastic) bottle.

         7.5.1  Sodium sulfate drying tubes — Plug the bottom of a large
                volume disposable pipet with a minimum amount of acidified
                glass wool  (Supelco Cat. No. 20383 or equivalent).   Fill the
                pipet halfway (3 g) with acidified sodium sulfate (See Sect.


    7.6  SULFURIC ACID — Reagent grade.

         7.6.1   Sulfuric  acid, 12 N -- Slowly add 335 mL concentrated sulfu-
                ric acid  to 665 mL of reagent water.

    7.7  SODIUM HYDROXIDE —.ACS  reagent grade or equivalent.

         7.7.1   Sodium hydroxide  IN — Dissolve  4.0 g reagent  grade sodium
                hydroxide in reagent water and dilute to 100 mL  in  volumetric
                flasks.

                                  515.2-9

-------
     7.7.2  Sodium hydroxide 6N
                                        i    '
7.8  ETHYL ETHER, UNPRESERVED — Nanograde, redistilled in glass if
     necessary.  Must be free of peroxides as indicated by EM Quant test
     strips (available from Scientific Products Co., Cat.  No. PI126-8,
     and other suppliers).  Procedures recommended for removal  of per-
     oxides are provided with the test strips.

7.9  ACIDIFIED SODIUM SULFATE — Cover 500 g sodium sulfate (Sect. 7.5)
     with ethyl ether (Sect. 7.8).  While agitating vigorously, add
     dropwise approximately 0.7 ml concentrated sulfuric acid.   Remove
     the ethyl ether overnight under vacuum and store the sodium sulfate
     in a 100°C oven.                   ]

7.10 CARBITOL, ACS GRADE — Available from Aldrich Chemical.

7.11 DIAZALD, ACS GRADE -- Available from Aldrich Chemical.

7.12 DIAZALD SOLUTION — Prepare a solution containing 10 g Diazald in
     100 mL of a 50:50 by volume mixture of ethyl ether and carbitol.
     This solution is stable for 1-month or longer when stored at 4°C in
     an amber bottle with a Teflon-lined screw cap.

7.13 TRIMETHYLSILYLDIAZOMETHANE (TMSD) — Available from Aldrich Chemical
     Co. as a 2 molar solution in hexane.  TMSD is stable during storage
     in this solution.                  t

7.14 4,4'-DIBROMOOCTAFLUOROBIPHENYL (DBOB) — 99% purity,  for use as
     internal standard.

7.15 2,4-DICHLOROPHENYLACETIC ACID (DCAA) — 99% purity, for use as
     surrogate standard.

7.16 POTASSIUM HYDROXIDE — ACS reagent grade or equivalent.

     7.16.1 Potassium hydroxide solution, 37% — Using extreme caution,
            dissolve 37 g reagent grade potassium hydroxide in reagent
            water and dilute to 100 mL.

7.17 STOCK STANDARD SOLUTIONS (1.00-2.00 /ig///L) — Stock standard solu-
     tions may be purchased as certified solutions or prepared from pure
     standard materials using the following procedure:

     7.17.1 Prepare stock standard solutions by accurately weighing
            approximately 0.0100-0.0200[g of pure material.  Dissolve the
            material in methanol and dilu-te to volume in a 10-mL volu-
            metric flask.  Larger volumes may be used at the convenience
            of the analyst.  If compound purity is certified at 96% or
            greater, the weight may be Used without correction to calcu-
            late the concentration of the stock standard.  Commercially
            prepared stock standards may be used at any concentration if


                              515.2-10

-------
             they are certified by the manufacturer or by an independent
             source.

      7.17.2  Transfer the stock standard solutions  into 15-mL TFE-fluoro-
             carbon-sealed screw cap amber vials.   Store at  4°C  or less
             when not in use.

      7.17.3  Stock standard solutions should  be  replaced after 2 months or
             sooner if comparison with laboratory fortified  blanks,  or QC
             samples  indicate  a problem.

      7.17.4  Primary  Dilution  Standards — Prepare  two sets  of standards
             according to the  sets labeled A  and B  in  Table  1.   For each
             set,  add approximately 25 ml  of  methanol  to a 50 ml volumet-
             ric  flask.   Add aliquots of each stock standard in  the range
             of approximately  20 to 400 pi and dilute  to volume  with
             methanol.   Individual  analyte concentrations will then be in
             the  range of 0.4  to 8 jug/mL (for a  1.0 mg/mL stock).   The
             minimum  concentration would be appropriate for  an analyte
             with  strong electron capture  detector  (ECD)  response,  e.g.
             pentachlorophenol.   The maximum  concentration is for  an
             analyte  with weak response, e.g., 2,4-DB.   The  concentrations
             given  in Table 2  reflect the  relative  volumes of stock stan-
             dards  used  for the  primary dilution standards used  in  gener-
             ating  the method  validation data.   Use these relative  values
             to determine the  aliquot volumes  of individual  stock  stan-
             dards  above.
   i
7.18  INTERNAL STANDARD  SOLUTION —  Prepare a  stock internal  standard
      solution by  accurately weighing approximately 0.050 g  of pure DBOB.
      Dissolve the  DBOB  in  methanol  and  dilute to volume  in  a 10-mL
      volumetric  flask.   Transfer the DBOB  solution  to  a  TFE-fluorocarbon-
      sealed  screw  cap bottle  and  store  at  room  temperature.  Prepare a
      primary dilution standard  at  approximately 1.00 jug/mL  by the  addi-
      tion of 20 Hi of the  stock standard  to  100 mL  of methanol.   Addition
      of 100 nl of  the primary dilution  standard solution to  the final 5
      mL of sample  extract  (Sect.  11)  results  in a  final  internal standard
      concentration of 0.020, Mg/mL.   Solution  should be replaced when
      ongoing QC  (Sect.  9)  indicates  a problem.  Note that DBOB has' been
      shown to be an  effective internal standard for the  method analytes,
      but other compounds may be  used  if the QC  requirements  in Sect. 9
      are met.

7.19  SURROGATE ANALYTE  SOLUTION —  Prepare a surrogate analyte stock
      standard solution  by  accurately weighing approximately  0.050 g of
     pure DCAA.   Dissolve the DCAA  in methanol and  dilute to volume in a
      10-mL volumetric flask.   Transfer the surrogate analyte solution to
      a TFE-fluorocarbon-sealed  screw cap bottle and store at room temper-
     ature.   Prepare a primary dilution standard at approximately 2.0
     Mg/mL by addition of 40 /*L at the stock standard to 100 mL  of
     methanol..  Addition of 250 fil of the surrogate analyte  solution to  a
     250-mL sample prior to extraction results in  a surrogate concentra-

                              515.2-11

-------
          tion  in the sample of 2 fig/I and,  assuming  quantitative  recovery  of
          DCAA,  a surrogate analyte concentration in  the final  5 ml  extract of
          0 1 ug/mL.   The surrogate standard solution should be replaced  when
          ongoing QC  (Sect. 9)  indicates a problem.   DCAA has been shown  to be
          an effective surrogate standard for the method analytes, but other
          compounds may be used if the QC requirements in Sect. 9  are met.

     7 20 INSTRUMENT  PERFORMANCE CHECK SOLUTION — Prepare a diluted dinoseb
          solution by adding 10 nl of the 1.0 [ig/nl dinoseb stock  solution  to
          the MTBE and diluting to volume in a 10-mL volumetric flask.  To
          prepare the check solution, add 40 ML of the diluted dinoseb solu-
          tion   16 U.L of the 4-nitrophenol stock solution, 6 /zL of the 3,5-
         'dichiorobenzoic acid stock solution, 50 /iL of the surrogate standard
          solution  25 pi of the internal standard solution, and  250 ML of
          methanol to a 5-mL volumetric flask and dilute to volume with MTBE.
          Methyl ate sample as described in Sect. 11.4.  Dilute the sample to
          10 ml in MTBE.  Transfer to a TFE-fluorocarbon-sealed screw cap
          bottle and store at room temperature.  Solution should be replaced
          when ongoing QC  (Sect. 9)  indicates a problem.

8.   SAMPLE COLLECTION. PRESERVATION. AND STORAGE

     8 1  Grab samples should be collected in 1-L amber glass  containers.
          Conventional sampling  practices (7) should  be followed;  however,  the
          bottle must not  be prerinsed with  sample before collection.

     8.2  SAMPLE  PRESERVATION AND  STORAGE  ,

          821   If  residual chlorine  is  present, add 80  mg  of sodium thiosul-
                  fate  (or  50 mg  of sodium sulfite) per  liter of sample to  the
                  sample  bottle  prior to  collecting the  sample.  Demonstration
                  data  in Section 17  of  this  method was  obtained using sodium
                  thiosulfate.

           822   After  the sample  is collected  in the bottle containing  the
                  dechlorinating  agent,  seal  the  bottle  and  mix to dissolve the
                  thiosulfate.

           8.2.3   Add hydrochloric  acid  (diluted  1:1  in  reagent water) to the
                  sample at the  sampling site in  amounts to  produce  a sample pH
                  < 2.   Short  range (0-3)  pH paper (Sect.  6.14) may  be used to
                  monitor the  pH.  Note:  Do  not  attempt  to mix  sodium thiosul-
                  fate and  HC1  in the sample bottle  prior to sample  collection.

           824  The samples  must be iced or refrigerated at 4°C  away from
                  light from the time of collection  until  extraction.  Preser-
                  vation study results indicate that  the sample analytes  (mea-
                  sured as  total  acid),  except 5-hydroxy-dicamba,  are stable in
                  water for 14 days when stored under these conditions (Tables
                  8 and 9).  The concentration of 5-hydroxydicamba is seriously
                  degraded over 14 days in a biologically active matrix.  How-
                  ever, analyte stability will very likely be affected by the

                                    515.2-12

-------
                 matrix; therefore, the analyst should verify that the preser-
                 vation technique  is applicable to the samples under study.

     8.3  EXTRACT STORAGE

          8.3.1  Extracts should be stored at 4°C or less away from light.
                 Preservation study results indicate that most analytes are
                 stable for 14 days (Tables 8 and 9); however, the analyst
                 should verify appropriate extract holding times applicable to
                 the samples under study.

9.  QUALITY CONTROL

     9.1  Minimum QC requirements are initial  demonstration of laboratory
          capability, determination of surrogate compound recoveries in each
          sample and blank,  monitoring internal  standard peak area or height
          in each sample and blank, analysis of laboratory reagent blanks,
          laboratory fortified matrices,  laboratory fortified blanks,  and QC
          samples.   A MDL for each analyte must also be determined. .

     9,2  LABORATORY REAGENT'BLANKS (LRB)  ~ Before processing any samples,
          the analyst must demonstrate that all  glassware and reagent  inter-
          ferences  are under control.   Each time a set of samples is extracted
          or reagents are changed,  a LRB  must  be analyzed.   If within  the
          retention, time window of any analyte the LRB produces a.peak that
          would prevent the  determination  of that  analyte,  determine the
          source of contamination and  eliminate  the interference before
          processing samples.

     9.3  INITIAL DEMONSTRATION OF  CAPABILITY

          9.3.1  Select a representative  fortified concentration (about 10 to
                 20 times MDL,  or  a mid-point in  the calibration range - see
                 Table 4) for each analyte.   Prepare a primary dilution
                 standard containing  each analyte at 1000  times  selected
                 concentration.  With  a  syringe,  add 250 /iL  of the concen-
                 trate to each  of  four to seven.250  mL aliquots  of reagent
                 water,  and analyze each  aliquot  according to  procedures
                 beginning  in  Sect. 11.

          9.3.2  For each analyte  the  recovery  value for all of  these  samples
                 must  fall  in  the  range of ±  40%   of the fortified concentra-
                 tion.  The RSD  of the measurements  must be  30%  or less.   For
                 compounds  failing  this criteria,  this  procedure must  be
                 repeated using  fresh  samples until  satisfactory performance
                 has been demonstrated for all  analytes.

          9.3.3   For each analyte, determine  the  MDL.   Prepare a minimum of 7
                 LFBs  at  a  low concentration.   Fortification concentration in
                 Table  2 may be  used as a  guide,  or  use  calibration data
                 obtained in Section 10 to estimate  a  concentration for  each
                 analyte that will produce a  peak with a 3-5 times signal  to

                                  515.2-13

-------
              noise  response.   Extract  and  analyze each  replicate accord-
              ing  to Sections  11  and  12.   It  is recommended that these
              LFBs be prepared  and  analyzed over  a period of several days,
              so that day  to day  variations are reflected in the precision
              measurement.  Calculate mean  recovery and  standard deviation
              for  each analyte. Use the equation  given in Section 13 to
              calculate the MDL.

     9.3.4    The  initial  demonstration bf  capability is used primarily to
              preclude a laboratory from analyzing unknown samples via a
              new, unfamiliar method prior  to obtaining  some experience
              with it.  As laboratory personnel gain experience with this
              method the quality  of data should improve  beyond those
              required here.

9.4  The analyst  is permitted  to modify GC columns, GC  conditions,
     concentration  techniques  (i.e., evaporation techniques), internal
     standard or surrogate compounds.  Each time such method modifica-
     tions are made,  the  analyst must repeat the procedures in Sect. 9.3.

9.5  ASSESSING SURROGATE  RECOVERY

     9.5.1    When surrogate recovery from a sample or a blank is <60% or
              > 140%,  check calculations to locate possible errors, forti-
              fying  solutions for degradation, contamination, and instru-
              ment performance.   If those steps do not reveal the cause of
              the problem, reanalyze theiextract.

     9.5.2    If a blank extract reanalysis fails the 60-140% recovery
              criteria, the problem must be identified and corrected
              before  continuing.

     9.5.3    If sample extract reanalysis meets the surrogate recovery
             'criteria, report only data for the reanalyzed extract.  If
              sample  extract continues to fail the recovery criteria,
             report  all data for that sample as suspect.

9.6  ASSESSING THE  INTERNAL STANDARD

     9.6.1   When using the internal  standard (IS) calibration procedure,
             the analyst must monitor the IS response (peak area or peak
             height) of all  samples during each analysis day.   The IS
             response for any sample chromatogram should not deviate from
             the daily calibration check standard IS response by more
             than 30%.

     9.6.2    If >30% deviation occurs with an individual extract,  opti-
             mize instrument performance and inject a second aliquot of
             that extract.
                              515.2-14

-------
             9.6.2.1   If the reinjected aliquot produces an acceptable
                       internal standard response, report results for
                       that aliquot.

             9.6.2.2   If a deviation of greater than 30% is obtained for
                       the reinjected extract, analysis of the samples
                       should be repeated beginning with Sect. 11, pro-
                       vided the sample is still available.   Otherwise,
                       report results obtained from the reinjected ex-
                       tract, but annotate as suspect.

     9.6.3   If consecutive samples fail the IS response acceptance
             criteria, immediately analyze a medium calibration standard.

             9.6.3.1   If the standard provides a response within 20% of
                       the .predicted value, then follow procedures item-
                       ized in Sect. 9.6.2 for each sample failing the IS
                       response criterion.

             9.6.3.2   If the check standard provides a response which
                       deviates more than 20% of the predicted value,
                       then the analyst must recalibrate as  specified in
                      /Sect. 10.

9.7  ASSESSING LABORATORY PERFORMANCE — LABORATORY FORTIFIED BLANK

     9.7.1   The laboratory must analyze at least one laboratory forti-
             fied blank (LFB) sample with every 20 samples or one per
             sample set (all samples extracted within a 24-hr period)
             whichever is greater.  The concentration of each analyte in
             the LFB should be approximately the same as in  Sect. 9.3.1.
             Calculate percent recovery (X,-).   If the recovery of any
             analyte falls outside the control limits (See Sect. 9.7.2),
           ..•  that analyte is judged out of control, and the  source of the
             problem should be identified and resolved before continuing
             analyses.

     9.7.2   Until  sufficient data become available, usually a minimum of
             results from 20 to 30 analyses, each laboratory should
             assess laboratory performance against the control limits in
             Sect.  9.3.2 that are derived from the data in Table 2.  When
             sufficient internal performance data become available, ,
             develop control limits from the mean percent recovery (X)
             and standard deviation (S) of the percent recovery.  These
             data are used to establish upper and lower control limits as
             follows:

                       UPPER CONTROL LIMIT  = X + 3S
                       LOWER CONTROL LIMIT  = X - 3S

             After each five to ten new recovery measurements, new con-
             trol limits should be calculated using only the most recent

                              515.2-15

-------
             20-30 data points.  These calculated control limits should
             not exceed those established  in Sect. 9.3.2.

     9.7.3   At least quarterly, analyze a QCS  (Sect. 3.13) from an
             outside source.

9.8  ASSESSING ANALYTE RECOVERY - LABORATORY FORTIFIED SAMPLE MATRIX

     9.8.1   Each laboratory must analyze  a LFM for 10% of the samples or
             one sample concentration per  set, whichever is greater.  The
             concentration should not be less then the background concen-
             tration of the sample selected for fortification.  Ideally,
             the concentration should be the same as that used for the
             laboratory fortified blank (Sect. 9.7).  Over time, samples
             from all routine sample sources should be fortified.

     9.8.2   Calculate the percent recovery, P, of the concentration for
             each analyte, after correcting the measured concentration,
             X, from the fortified sample  for the background concentra-
             tion, b, measured in the unfortified sample.

             P = 100 (X - b) / fortified concentration,

             and compare these values to. control limits appropriate for
             reagent water data collected  in the same fashion.  Accep-
             tance criteria are the same as those in Section 9.7 for
             LFBs.

     9.8.3   If the recovery of any such analyte falls outside the desig-
             nated range, and the laboratory performance for that analyte
             is shown to be in control (Sect. 9.7), the recovery problem
             encountered with the fortified sample is judged to be matrix
             related, not system related.  The result for that analyte in
             the unfortified sample is labeled suspect/matrix to inform
             the data user that the results are suspect due to matrix
             effects.

9.9   ASSESSING INSTRUMENT SYSTEM/INSTRUMENT PERFORMANCE CHECK  (IPC)
      SAMPLE — Instrument performance should be monitored on a daily
      basis by analysis of the IPC sample.  The IPC sample contains
      compounds designed to monitor instrument sensitivity, column
      performance (primary column) and ehromatographic performance.  IPC
      sample components and performance criteria are listed in Table 11.
      The sensitivity requirements are set based on the MDLs published in
      this method.  If the laboratory MDLs differ from those demonstrated
      here, the amount of dinoseb in the IPC sample should be adjusted
      accordingly.

9.10  The laboratory may adopt additional QC practices for use with this
      method.  The specific practices that are most productive depend
      upon the needs of the laboratory and the nature of the samples.
      For example, field or laboratory duplicates may be analyzed to
      assess the precision of the envirbnmental measurements or field

                              515.2-16

-------
           reagent blanks may be used to assess contamination of samples under
           site conditions, transportation, and storage.

10.   CALIBRATION AND STANDARDIZATION

     10.1 Establish GC operating parameters equivalent to those indicated in
          Sect. 6.12.  'This calibration procedure employs procedural calibra-
          tion standards, i.e.,  fortified aqueous standards which are pro-
          cessed through the method (Sect.  11).  The GC system is calibrated
          by means of the internal  standard technique (Sect. 10.2).   NOTE:
          Calibration standard solutions must be prepared such that  no unre-
          solved analytes are mixed together (See Table 1).

     10.2 INTERNAL STANDARD CALIBRATION PROCEDURE — To use this approach, the
          analyst must select one or more internal  standards compatible in
          analytical  behavior to the compounds of interest.  The analyst must
          further demonstrate that  the measurement  of the internal  standard is
          not  affected by method or matrix  interferences.   DBOB (Sect.  7.14)
          has  been identified as a  suitable internal  standard.

          10.2.1  Prepare aqueous calibration standards at a minimum  of three
                 (five are recommended)  concentration levels for each method
                 analyte as  follows:  for each  concentration,  fill a  250-mL
                 volumetric  flask with 240  mL  of reagent  water  at pH 1  and
                 containing  50 g of dissolved  sodium  sulfate.   Add an appro-
                 priate aliquot  of  the primary dilution  standard (Sect.  7.17-
                 .4)  and dilute  to  250 mL with the  same  reagent  water.   Guid-
                 ance  on the  number of standards  is  as follows:  A minimum of
                 three calibration  standards  are  required  to  calibrate  a  range
                 of a  factor  of  20  in  concentration.   For  a factor of 50  use
                 at least  four standards, and  for a  factor of 100 at least  .-
                 five  standards.  The  lowest standard  should  represent  analyte
                 concentrations  near,  but above,  their respective MDLs.   The
                 remaining standards should  bracket the analyte  concentrations
                 expected  in  the  sample  extracts, or  should define the working
                 range  of  the  detector.   Process each aqueous calibration
                 sample  through  the analytical  procedure beginning with Sect.
                 11.1.2.  The  internal standard is added to the  final 5 mL
                 extract as specified  in Sect.  11.4.3  or 11.5.9.

         10.2.2 Analyze each  calibration standard according to the  procedure
                beginning in  Sect.  11.1.2.   Tabulate  response (peak  height or
                area) against concentration for each  compound and internal
                standard.  Calculate the response factor  (RF) for each anal-
                .yte and surrogate using Equation 1.
                 RF =
(As)  (C,.)

(Ai.) (C.)
Equation 1
                                  515.2-17

-------
                 where:

                 As  = Response for the analyte to be measured.
                 Ais  = Response for the internal  standard.
                 Cis  = Concentration of the internal  standard  (/zg/L).
                 Cs  = Concentration of the analyte to be measured (/jg/L).

          10.2.3 If the RF value over the working range is constant (30%  RSD
                 or less) the  average RF can be used for calculations.  Alter-
                 natively, the results can be used to plot a calibration  curve
                 of response ratios  (A /Ais) vs. Cs.  A data station may be
                 used to collect the cnromatographic data, calculate response
                 factors and generate linear or second order regression
                 curves.

          10.2.4 The working calibration curve or RF must be verified on  each
                 working shift (not  to exceed 12 hours) by the measurement of
                 one or more calibration standards.  It is highly recommended
                 that a calibration  verification be performed  at the beginning
                 and at the end of  every extended period of instrument opera-
                 tion so that  field  sample extracts are bracketed by calibra-
                 tion standards.   It is also recommended that  more that one
                 standard concentration be analyzed.so that the calibration is
                 verified at more than one point.  New calibration standards
                 need not be derivatized each day.  The same standard extract
                 can be used up to  14 days.  If the response for any analyte
                 varies from the predicted response by more than +30%, the
                 test must be  repeated using a fresh calibration standard.  If
                 the repetition also fails; a new calibration  curve must  be
                 generated for that  analyte using freshly prepared standards.
                 For those analytes  that failed the calibration verification,
                 results from  field  samples analyzed since the last passing
                 calibration should  be considered suspect.  Reanalyze sample
                 extracts for  these  analytes after acceptable  calibration is
                 restored.

          10.2.5 Verify calibration  standards periodically, at least quarterly
                 is recommended, by  analyzing a standard prepared from refer-
                 ence material obtained from an independent source.  Results
                 from these analyses must be within the limits used to rou-
                 tinely check  calibration.

11.   PROCEDURE

     11.1 MANUAL HYDROLYSIS AND SEPARATION OF INTERFERENCES

          11.1.1 Remove the sample  bottles from cold storage and allow them to
                 equilibrate to room temperature.  Acidify and add sodium
                 thiosulfate to LFBs, LRBs and QCSs as specified in Sect. 8.

          11.1.2 Measure a 250-mL aliquot of each sample with  a 250-tnL gradu-
                 ated cylinder and  pour into a 500-mL separatory funnel.  Add
                 250 /iL of the surrogate primary dilution standard (Sect.

                                    515.2-18

-------
            7.19) to each 250-mL sample.  The surrogate will be at a
            concentration of 2 /ig/L. Dissolve 50 g sodium sulfate in the
            sample.

     11.1.3 Add 4 ml of 6 N NaOH to each sample, seal, and shake.  Check
            the pH of the sample with pH paper or a pH meter; if the
            sample does not have a pH greater than or equal  to 12, adjust
            the pH by adding more 6 N NaOH.  Let the sample sit at room
            temperature for 1 hr, shaking the separatory funnel and
            contents periodically.  Note: Since many of the herbicides
            contained in this method are applied as a variety of esters
            and salts, it is vital to hydrolyze them to the parent acid
            prior to extraction.  This step must be included in the
            analysis of all extracted field samples, LRBs, LFBs, LFMs,
            and QCS.

     11.1.4 Add 15 ml methylene chloride to the graduated cylinder to
            rinse the walls, transfer the methylene chloride to the
            separatory funnel and extract the sample by vigorously shak-
            ing the funnel for 2 min with periodic venting to release
            excess pressure.  Allow the organic layer to separate from
            the water phase for a minimum of 10 min.  If the emulsion
            interface between layers is more than one-third the volume of
            the solvent layer, the analyst must employ mechanical tech-
            niques to complete the phase separation.  The optimum tech-
            nique depends upon the sample, but may include stirring,
            filtration through glass wool, centrifugation, or other
            physical methods.  Discard the methylene chloride phase
            (Sect.14,15).

     11.1.5 Add a second 15-mL volume of methylene chloride to the separ-
            atory funnel and repeat the extraction procedure a second
            time, discarding the methylene chloride layer.  Perform a
            third extraction in the same manner.

     11.1.6 Drain the contents of the separatory funnel  into a 500-mL
            beaker.  Adjust the pH to 1.0 ± 0.1 by the dropwise addition
            of concentrated sulfuric acid with constant stirring. Monitor
            the pH with a pH meter (Sect. 6.8) or short range (0-3) pH
            paper (Sect. 6.14).

11.2 SAMPLE EXTRACTION

     11.2.1 Vacuum Manifold — Assemble a manifold (Sect. 6.3) consisting
            of 6-8 vacuum flasks with filter funnels (Sect.  6.1,6.2).
            Individual vacuum control, on-off and vacuum release valves
            and vacuum gauges are desirable.  Place the 47 mm extraction
            disks (Sect. 7.1) on the filter frits.

     11.2.2 Add 20 mL of 10% by volume of methanol in MTBE to the top of
            each disk without vacuum and allow the solvent to remain for
            2 min.  Turn on full vacuum and draw the solvent through the
            disks, followed by room air for 5 min.

                              515.2-19

-------
     11.2.3 Adjust the vacuum to approximately 5 in.  (mercury)  and  add
            the following in series to the filter funnel  (a)  20 ml
            methanol  (b) 20 mL reagent wa;ter (c) sample.   Do  not allow
            the disk to dry between steps and maintain the vacuum at 5
            in.

     11.2.4 After all the sample has passed through the disk,  apply
            maximum vacuum and draw room air through  the disks  for  20
            min.

     11.2.5 Place the culture tubes (Sect. 6.4) in the vacuum tubes to
            collect the eluates.  Elute t\\e disks with two each 2-mL
            aliquots of 10% methanol in MTBE.  Allow each aliquot to
            remain on the disk for one mih before applying vacuum.

     11.2.6 Rinse each 500-mL beaker (Sect.ll;1.6) with 4 mL  of pure MTBE
            and elute the disk with this solvent as in Sect.  11.2.5.

     11.2.7 Remove the culture tubes and cap.
                                         j'
11.3 EXTRACT PREPARATION                 ;

     11.3.1 Pre-rinse the drying tubes (Sect. 7.5.1)  with 2 mL of MTBE.

     11.3.2 Remove the entire extract with a 5-mL pipet and drain the
            lower aqueous layer back into the culture tube.  Add the
            organic layer to the sodium sulfate drying tube (Sect.
            7.5.1).  Maintain liquid in the drying tube between this and
            subsequent steps.  Collect the dried extract in a 15-mL
            graduated centrifuge tube or a 10-mL Kuderna-Danish tube.

     11.3.3 Rinse the culture tube with an additional 1 mL of MTBE and
            repeat Sect. 11.3.2.         :

     11.3.4 Repeat step Sect. 11.3.3 and finally add a 1-mL aliquot of
            MTBE to the drying tube before it empties.  The final volume
            should be 6-9 mL.  In this form the extract is esterified as
            described below.

11.4 EXTRACT ESTERIFICATION WITH DIAZOMETjHANE — See Section 11.5 for
     alternative procedure.              j

     11.4.1 Assemble the diazomethane generator (Figure 1) in a hood.

     11.4.2 Add 5 mL of ethyl ether to Tube 1.  Add 4 mL of Diazald
            solution (Sect. 7.12) and 3 mL of 37% KOH solution  (Sect.
            7.16.1) to the reaction tube 2.  Immediately place the exit
            tube into the collection tube containing the sample extract.
            Apply nitrogen flow (10 mL/min) to bubble diazo-methane
            through the extract.  Each charge of the generator should be
            sufficient to esterify four samples.  The appearance of a
            persistent yellow color is an indication that esterification
            is complete.  The first sample should require 30 sec to 1 min

                              515.2-20

-------
            and each subsequent sample somewhat longer.  The final sample
            may require 2-3 mi IT.

     11.4.3 Cap each collection tube and allow to remain stored at room
            temperature in a hood for 30 min.  No significant fading of
            the yellow color should occur during this period. Fortify
            each sample with 100 juL of the internal standard primary
            dilution solution (Sect. 7.18) and reduce the volume to 5.0
            ml with the analytical concentrator (Sect. 6.10), a stream of
            dry nitrogen,  or an equivalent concentration technique.
            NOTE:   The excess diazomethane is volatilized from the
            extract during the concentration procedure.

     11.4.4 Cap the tubes  and store in a refrigerator if further process-
            ing will not be performed immediately.   Analyze by GC-ECD.

11.5 EXTRACT ESTERIFICATION WITH TRIMETHYLSILYLDIAZOMETHANE (TMSD)  --
    •'Alternative .procedure.  It should be noted that the gas
     chromatographic background is significantly increased when TMSD is'
     used as the derivatizing reagent instead of the generated
     diazomethane.   Although no method analyte is affected by this
     increased background,  the recommended surrogate,  2,4-dichloro-
     phenylacetic  acid,  is  masked by an interfering peak.   This renders
     the surrogate  useless  at 1 jiig/L or lower.  Any compound found
     suitable  when  TMSD is  used is acceptable as a  surrogate.

     11.5.1 Carry  out the  hydrolysis,  clean-up,  and extraction of the
            method  analytes as described up to Sect.  11.2.4.

     11.5.2 Elute  the herbicides from the disk by passing  two 2 ml
            aliquots of methyl  tertiary butyl  ether (MTBE)  through the
            disk into the  collection tube.   Rinse the  sample container
            with 4  ml of MTBE and pass  it through the  disk into the tube.

     11.5.3 Transfer the MTBE extract  from the collection  tube into an
            anhydrous sodium sulfate drying tube which has  been pre-
            wetted  with  1 ml MTBE.   Be  sure to discard any  water layer.

     11.5.4 Before  the  extract  passes  completely through the sodium
            sulfate,  add an additional  2  ml of MTBE  as a rinse.

     11.5.5 Concentrate  the dried  extract to  approximately  4 ml.   Add
            methanol  (approx.  1  ml)  to  the  extract  to  yield  a  20%  (v/v)
            methanol  in  MTBE solution.  Adjust the  volume  to 5  ml  with
            MTBE.   (TMSD produces  the most  efficient methylation of the
            herbicides  in a 20%  methanol,  80%  MTBE  solution.)

     11.5.6 Add  50  fil of the  2  M TMSD solution to each 5 mL  sample
            extract.

     11.5.7 Place the tube  containing the extract into a heating block at
            50°C and  heat the extract for 1 hour.
                             515.2-21

-------
     11.5.8 Allow the extract to cool to room temperature, then add 100
            IJ.L of 2 M acetic acid in mei;hanol to react with any excess
            TMSD.

     11.5.9 Fortify the extract with 100 ill of the internal standard
            solution (See Sect. 7.18) to yield a concentration of 0.020
            tig/ml.

    11.2.10 Proceed with the identification and measurement of the
            analytes using GC/ECD according to the procedures described
            in Sect 11.6.

11.6 GAS CHROMATOGRAPHY

     11.6.1 Sect. 6.12 summarizes the recommended GC operating
            conditions.  Included in Table 1 are retention times observed
            using this method.  Figures 2A and 2B illustrate the
            chromatographic performance of the primary column (Sect.
            6.12.1) for groups A and B pf the method analytes.  Other GC
            columns or chromatographic Conditions, may be used if the
            requirements of Sect. 9.3 are met.

     11.6.2 Calibrate or verify the existing calibration daily as
            described in Sect. 10.
     11.6.3 Inject 2 /*L of the sample extract.
            size in area units.
Record the resulting peak
     11.6.4 If the response for any sample peak exceeds the working range
            of the detector, dilute the extract and reanalyze.  Add
            additional IS, so that the ;IS amount in the extract will be
            the same as in the calibration standards.

11.7 IDENTIFICATION OF ANALYTES

     11.7.1 Identify a sample component by comparison of its retention
            time to the retention time of a reference chromatogram.  If
            the retention time of an unknown compound corresponds, within
            limits, to the retention time of a standard compound, then an
            analyte is considered to be identified.

     11.7.2 The width of the retention time window used to make identi-
            fications should be based upon measurements of actual
            retention time variations of standards over the course of a
            day.  Three times the standard deviation of a retention time
            can be used to calculate a suggested window size for  a
            compound.  However, the experience of the analyst  should
            weigh heavily in interpretation of chromatograms.

     11.7.3 Identification requires expert judgment when sample compo-
            nents are not resolved chromatographically.  When  GC  peaks
            obviously represent more than one sample component (i.e.,
            broadened peak with shoulder(s) or valley between  two or more

                              515.2-22

-------
    ;   •-'•-.       maxima, ,or any time doubt exists over  the  identification  of a
                  peak in a chromatogram, appropriate alternative  techniques  to
                  help confirm peak identification need  to be employed.   For
                  example, more positive identification  may  be made  by the  use
       >•:•          of an alternative detector which operates  on a
    ,,   .     ...-.V;  chemical/physical principle different  from that  originally
                  used, e.g., mass spectrometry, or the  use  of a second
                  chromatography column.  A suggested alternative  column  is
       -,;.••  .- ,    described in Sect, 6.12.2.              .-.-.-.

 12.   DATA ANALYSIS AND CALCULATIONS

      12.1 Calculate analyte concentrations in the sample from the response for
           the  analyte using the calibration procedure described in  Sect. 10.
           Use.,the multi-point calibration for each analyte  to make  all
           calculations.   Do not use the daily calibration verification data  to
           quantitate analytes in samples.

      12.2 Calculate the  concentration (C)  in  the sample using the response
           factor  (RF)  determined in Sect.  10.2.2 and Equation 2,  or determine
           sample  concentration from the calibration  curve-(Sect.  10.2.3).

               ' '    ,  ..-.;    '  ;•    CAS)(IS)          .   ..*
                  c  (M9/L)  =                  .  ..     Equation 2.
                                  (Ais)(RF)(V0)

          where:

                  As  .== Response, for the analyte.to be measured.
                  Ajs = Response for the internal standard. •  ' '.
                  I.s  .= Amount of internal standard added to  each
                       ,   extract  (/jg),.- • ,             ,
                  V0  = Volume of water extracted (L).

13.   METHOD PERFORMANCE

     13.L All data .shown in Section 17 of this  method were obtained  with  the
          diazomethane-esterification option.

     13.2 In a single laboratory,  analyte recoveries from reagent  water were
          determined at three concentration levels, Tables 2-4.  Results  were
          used to determine,thfi analyte MDLs  (8) listed  in Table 2.  The
          calculation, fordetermining MDL is:
                 where^,'?S t(n"1/1"alpha = °-99).
                          '^(n-i.i-aipha = o.99)  = Student's  t value for the 99%

                          confidence level with n-1 degrees of freedom

                          n - number of replicates

                          S = standard deviation,of replicate analyses.

                                   515.2-23

-------
13.3 In a single laboratory, analyte recoveries from dechlorinated tap
     water were determined at two concentrations,  Tables 5 and 6.   In
     addition, analyte recoveries were determined at two concentrations
     from an ozonated surface (river) water, Tables 7 and 8,  and at one
     level from a high humectant surface (reservoir) water,  Table 10.
     Finally, a holding study was conducted on the preserved,  ozonated
     surface water and recovery data are presented for day 1  and day 14
     of this study, Tables 8 and 9.  The ozonated surface water was
     chosen as the matrix in which to study analyte stability during a
     14-day holding time because it was very biologically active.
                                                                     This
14.  POLLUTION PREVENTION

     14.1 This method utilizes liquid-solid extraction technology which
          requires the use of very small quantities of organic solvents.
          feature eliminates the hazards involved with the use of large
          volumes of potentially harmful organic solvents needed for
          conventional liquid-liquid extractions.  Also, mercuric chloride, a
          highly toxic and environmentally hazardous chemical, has been
          replaced with hydrochloric acid as the sample preservative.  These
          features make this method much safer and a great deal less harmful
          to the environment.

     14.2 For information about pollution prevention that may be applicable to
          laboratory operations, consult "Less is Better:  Laboratory Chemical
          Management for Waste Reduction" available from the American Chemical
          Society's Department of Government Relations and Science Policy,
          1155 16th Street N.W., Washington, D.C. 20036.

15.  WASTE MANAGEMENT

     15.1 Due to the nature of this method, there is little need for waste
          management.  No large volumes of solvents or hazardous chemicals are
          used.  The matrices of concern are finished drinking water or source
          water.  However, the Agency requires that laboratory waste manage-
          ment practices be conducted consistent with all applicable rules and
          regulations, and that laboratories protect the air, water, and land
          by minimizing and controlling all releases from fume hoods and bench
          operations.  Also, compliance is required with any  sewage discharge
          permits and regulations, particularly the hazardous waste identifi-
          cation rules and land disposal restrictions.  For further informa-
          tion on waste management, consult "The Waste Management Manual for
          Laboratory Personnel," also available from the American Chemical
          Society at the address in Sect. 14.2.

16.  REFERENCES

     1.   ASTM Annual Book of Standards, Part  11, Volume 11.02, D3694-82,
          "Standard Practice for Preparatioji of Sample Containers and  for
          Preservation," American Society for  Testing and Materials,
          Philadelphia, PA, p. 86, 1986.
                               515.2-24

-------
 2'  •Sln?; Committee on Chemical  Safety, 3rd Edition,



 7.   ASTM Annual Book  of  Standards, Part 11, Volume 11.01,  D3370-82
      Standard Practice for Sampling Water," American Society for Testing
     and Materials, Philadelphia, PA, p. 130, 1986.               '"iing
           ,  J.A., Foerst, D.L., McKee, G.D., Quave, S.A., and Budde,

     ii!" 1426-1435    yS6S     Wastewaters," Environ. Sen. Techno! .  1981,


9.   40 CFR,  Part 136, Appendix B.
                             515.2-25

-------
17.  TABLES.  DIAGRAMS. FLOWCHARTS AND VALIDATION DATA
                             TABLE 1.   RETENTION DATA
Retention Time,
Analvte Groirn3 Primarv
3,5-Dichlorobenzoic acid
2,4-Dichlorophenylacetic acid (SA)
Dicamba
Dichlorprop
2,4-D
4,4'-Dibromooctafluorobiphenyl (IS)
Pentachlorophenol
Si 1 vex
5-Hydroxydicamba
2,4,5-T
2,4-DB
Dinoseb
Bentazon
Picloram
Dacthal diacid metabolite
Acifluorfen
A ' ;
A,B
B
A
B
A,B
A
B
B ,
A '
B
A
B
B
A
B
16.72
19.78
20.18
22.53
23.13
24.26
25.03
25.82
26.28
26.57
27.95
28.03
28.70
29.93
31.02
35.62
min.D
Confirmation
18.98
22.83
23.42
25.90
27.01
26.57
27.23
29.08
30.18
30.33
31.47
33.02
33.58
35.90
34.32
40.58
 a  Analytes were  divided  into  two  groups during method development to
   avoid  chromatographic  overlap.

 b  Columns and  chromatographic conditions  are described  in Sect. 6.12.
                                      515.2-26

-------
                TABLE 2.   SINGLE LABORATORY RECOVERY,  PRECISION DATA
                          AND METHOD DETECTION LIMIT WITH FORTIFIED
                          REAGENT WATER - LEVEL 1
Analvte
Acifluorfen
Bentazon
2,4-D
2,4-DB
Dacthal diacid metabolite
Dicafnba
3,5-Dichlorobenzoic acid
Dichlorprop
Dinoseb
5-Hydroxydicamba
Pentachlorophenol
Picloram
2,4,5-T
2,4,5-TP
Fortified
Cone.
UQ/L
0.50
2.50
0.25
2.50
0.25
1 0.75
1.25
0.25
0.50
0.75
0.25
0.75
0.25
0.25
Mean3
Recovery
%
78
70
96
79
96
109
126
106
87
90
103
95
116
98
Relative
Std. Dev.
%
21
11
38
12
16
11
24
15
22
12
18
15
18
9
MDL
ua/L
0.25
0.63
0.28
0.72
0.13 .
0.28
1.23
0.13
0.28
0.25
0.16
0.35
0.16
0.06
a  Based on the analyses of seven replicates.
                                     515.2-27

-------
             TABLE 3.  SINGLE LABORATORY RECOVERY AND PRECISION DATA
                        FOR FORTIFIED  REAGENT WATER -  LEVEL 2
Anal vte 	 . 	
Acifluorfen
Bentazon
2,4-D
2,4-DB
Dacthal diacid metabolite
Dicamba
3,5-Dichlorobenzoic acid
Dichlorprop
Dinoseb
5-Hydroxydicamba
Pentachlorophenol
Picloram
2,4,5-T
2,4,5-TP
Fortified
Cone.
ua/l
0.80
4.0
0.40
4.0
0.40
I
1.20
2.00
0.40
0.80
1.20
0.40
1.20
0.40
0.40
Mean3
Recovery
%
61
81
96
90
96
109
126
76
87
90
66
68
116
105
Relative
Std. Dev.
%
27
8
38
13
16
11
24
21
22
12
26
21
18
7
a  Based on the analyses of six-seven replicates.
                                     515.2-28

-------
               TABLE 4.   SINGLE LABORATORY RECOVERY AND PRECISION DATA
                         FOR FORTIFIED REAGENT WATER - LEVEL 3
Analvte
Acifluorfen
Bentazon
2,4-D
2,4-DB 	
Dacthal diacid metabolite
Dicamba
3,5-Dichlorobenzoic acid
Dichlorprop
Dinoseb
5-Hydroxydicamba
Pentachlorophenol
Picloram
2,4,5-T
2,4,5-TP
Fortified
Cone.
ua/L
2.0
10.0
1.0
10.0
1.0
3.0
5.0
1.0
2.0
3.0
1.0
3.0
1.0
1.0
Mean3
Recovery
%
59
68
90
74
60
75
62
97
63
77
69
66
64
68
Relative
-Std. Dev.
01
13
8
20
,6
10
9
18
17
10
8
11
9
15
8
a  Based on the analyses of six-seven replicates.
                                     515.2-29

-------
             TABLE 5.  SINGLE LABORATORY RECOVERY AND PRECISION DATA
                        FOR FORTIFIED,  DECHLORINATED  TAP WATER -  LEVEL  1
Analvte
Acifluorfen
Bentazon
2,4-D
2,4-DB
Dacthal diacid metabolite
Dicamba
3,5-Dichlorobenzoic acid
Dichlorprop
Dinoseb
5-Hydroxydicamba
Pentachl orophenol
Picloram
2,4,5-T
2,4,5-TP
Fortified
Cone.
' aa/L
0.50
2.50
0.25
2.50
0.25
0.75
1.25
0.25
0.50
0.75
0.25
i
0.75
0.25
0.25
Mean3
Recovery
%
117
96
59b
112
101
91
103
218C
134
90
91
76
118
99
Relative
Std. Dev.
%
21
12
55
15
10
14
15
37
'10
14
8
28
16
10
a  Based on the analyses of six-seven replicates.
b  2,4-D background value was 0.29 /ig/L.
c  Probable interference.
                                     515.2-30

-------
              TABLE  6.  SINGLE  LABORATORY RECOVERY AND PRECISION DATA'
                         FOR FORTIFIED,  DECHLORINATED TAP WATER - LEVEL  2
Analvte
Acifluorfen
Bentazon
2,4-D
2,4-DB
Dacthal diacid metabolite
Dicamba
3,5-Dichlorobenzoic acid
Dichlorprop
Dinoseb
5-Hydroxydicamba
Pentachlorophenol
Picloram
2,4,5-T
2,4,5-TP
2,4-Dichlorophenylacetic acid6
Fortified
Cone.
ua/L
2.0
10.0
1.0
10.0
1.0
3.0
5.0
1.0
2.0
3.0
1.0
3.0
1.0
1.0
1.0
Mean3
Recovery
%
150
112
90
111
118
86
111
88
121
96
96
132
108
115
120
Relative
Std. Dev.
"/
7
9
16
10
8
10
5
30
6
6
6
12
10
7
19
a  Based on the analyses of six-seven replicates.

b  Surrogate analyte.
                                     515.2-31

-------
              TABLE 7.   SINGLE LABORATORY  RECOVERY  AND  PRECISION DATA
                        FOR FORTIFIED, OZONATED SURFACE WATER - LEVEL 1
Analvte
Acifluorfen
Bentazon
2,4-D
2,4-DB
Dacthal diacid metabolite
Dicamba
3,5-Dichlorobenzoic acid
Dichlorprop
Dinoseb
5-Hydroxydicamba
Pentachl orophenol
Picloram
2,4,5-T
2,4,5-TP
2,4-Dichlorophenylacetic acidb
Fortified
Cone.
uq/L
0.50
2.50
0.25
2.50
0.25
0.75
1.25 '
0.25
\
0.50
0.75
0.25
0.75
0.25
0.25
0.25
Mean3
Recovery
%
172
92
127
154
113
107
100
115
134
89
110
109
102
127
72
Relative
Std. Dev.
%
14
22
13
19
17
13
17
20
28
13
22
27
19
8
31
a  Based on the analyses of six-seven replicates.

b  Surrogate analyte.
                                     515.2-32

-------
       TABLE 8.  SINGLE LABORATORY RECOVERY AND PRECISION DATA FOR FORTIFIED
                 OZONATED SURFACE WATER - LEVEL 2, STABILITY STUDY DAY lb   '
Analvte
Acifluorfen
Bentazon
2,4-D
2,4-DB
Dacthal diacid metabolite
Dicamba
3,5-Dichlorobenzoic acid
Dichlorprop
Dinoseb
5-Hydroxydicatnba
Pentachlorophenol
Picloram
2,4,5-T
2,4,5-TP
2,4-DichTorophenylacetic acidc
Fortified
Cone.
UQ/l
2.0
10.0
1.0
10.0
1.0
3.0
5.0
1.0
2.0
3.0
1.0
3.0
1.0
1.0
1.0
Mean3
Recovery
%
173
122
126
130
116
109
115
116
116
121
118
182
112
122
110
Relative
Std. Dev.
°/
11
7
10
7
11
9
. 11
11
9
9
10
14
9
10
26
a  Based on the analyses -of six-seven replicates.
b  Samples preserved at pH = 2.0.
c  Surrogate analyte.
                                     515.2-33

-------
TABLE 9. SINGLE LABORATORY RECOVERY AND PRECISION DATA FOR FORTIFIED,
OZONATED SURFACE WATER - LEVEL 2, STABILITY STUDY DAY 14b
Analvte
Acifluorfen
Bentazon
2,4-D
2,4-DB
Dacthal diacid metabolite
Dicamba
3,5-Dichlorobenzoic acid
Dichlorprop
Dinoseb
5-Hydroxydicamba
Pentachlorophenol
Picloram
2,4,5-T
2,4,5-TP
2,4-Dichlorophenylacetic acidc
Fortified
Cone.
ULQ/l
2.0
10.0"
1.0
10.0
1.0
3.0
5.0
i
1.0
2.0
3.0
1.0 -
3.0 :
1.0
1.0
1.0 '.
Mean3
Recovery
% •
151
97
84
128
116 •
103
81
107
118
20
94
110
113
113
87
Relative
Std. Dev. ,
• %
18
9
11
10
7
9
'• ' ••'•'•. T2 '" ' :;"
11 '
7 .
;' '14 '
' -1' 7
32
8
11
... 6
a  Based on the analyses of six-seven replicates.
b  Samples preserved at pH = 2.0.
c  Surrogate analyte.
                                     515.2-34

-------
             TABLE 10.   SINGLE LABORATORY RECOVERY AND PRECISION DATA FOR
                        FORTIFIED,  HIGH HUMIC  CONTENT  SURFACE  WATER
Analvte
Acifluorfen
Bentazon
2,4-D
2,4-DB
Dacthal diacid metabolite
Dicamba
3,5-Dichlorobenzoic acid
Dichlorprop
Dinoseb
5-Hydroxydicamba
Pentachlorophenol
Picloram
2,4,5-T
2,4,5-TP
Fortified
Cone.
//a/I
2.0
10.0
1.0
10.0
1.0
3.0
5.0
1.0
2.0
3.0
1.0
3.0
1.0
1.0
Mean3
Recovery
%
120
87
59
80
100
76
87
110
97
82
70
124
101
80

Relative
Std. Dev.
"/
13
11
7
14
6
9
4
22
6
9
5
9
4
6
a  Based on the analyses of six-seven replicates.
                                     515.2-35

-------




















z
o
1—

o
V)

o
LU
1C

LU
W

S;
0£
O
U.
LU
O.

?^™
O
S
o
CO
«c
_1
I— 1
LU
O3
£
1—























CO
+i
01

03
C
03

o

c
o
'£
cu
4J
CU
O



^"
O
o
0








cu
co
o
c
Q









£>
'>
'^
CO
CU
co














XI
LO
o
1 — I
V
u_
CJ3
V
o

o




to
1 — 1







o
cu
CL
o
-t->
«=»•

O)
u
03
E

o
S-
cu
Q.
O
t—
a.
03
cn
o
< ^
03
0
,—
O













JD
5
•
O
A

E
O
3
0
CO
cu
o:




to to
O r-(


•a
o
03
0
•r—
O
M
C
a>
o "o
s- c
0 CU
[c Q.
o o
Q -M
LO "Z.
oo «s-






cu
o
s-
o
s-
cu
a.
c:
1
o
C_)
] t
03
t
co
1 O
; cu
01

C

J-
•o
'3

OS
cu
a.
a>
•*->
d
•r—
O
~^_
I — 1

-*
c ^
-!° «

3 • °°
8- I
w c
T^ """
•f-*
'*'-'* f—
^ en
3 «
2 -
"s -^
03
•a
tl '3
i3 cu .*:
^ r-H ^— CU
u_ — O Q.
3 i-H
f~ "^- CU
ro Xr-t, ^
oo 00 V) -=
3 • -r- cn
5 rt — 'S
II OJ -C

i-nj U- r-H .C
CO ^3 " — +J
Q" CU
u_ cu cu
g 3§

CO



























• "
o
03
3
cr
cu
cu
^

^*

•a
cu
c
<£
cu
-a
01
03
CO
03
CU
a.
o
^»

cu
^J

c
cu
cu
3
O)
,-Q
C
O
"43
= ^
00
(S C

ja
OS
cu
a.

cu
cn
03

cu
^>
03

CU
5
CO


•o
C
o3
co
ca
cu
CL

O
CU
.c:

^
O)
+j

-^
oo"
CU CO
CO *^
03
£= CU
•r- Q.
00 O
cu 3
E +->
4-> CU
C ^->
o-
•i- U-
•4-> O
3
r— CU
CU C
•i — CU
CO
CU 03
o -a
cu cu
S_ -C
CU 4->
r i
' I
•i— 03
-o
CU •
.C 00
+J O
CU
co co
"r*"
c.
+-> -r-
CU -C
• s- +J
I cu -o
_^^ *r~


515.2^36

-------
          HOW
                    4- f\AT JOINT WJTH 0 WHO AND CUM*
OIETHYL ETHERIEVU
                             f LAT XXNT WTH 0 JUNG AND CUM*
            OlAZALOUVfL
               KOH
     FIGURE 1. DIAZOMETHANE GENERATOR
                         515.2-37

-------
 i
S
O
O
 •
n
o
o
 •
                      X


                    515.2-38


§
                                           (0
                                O  ««

                               '".  X
                                                 2
                                      ««
                                      >,§.
                                      ** v'
                                                 3 «•»
                                                 P «
                                                 »- :*
                                                 III

                                                 -C N ,
                                                 oo
                                            t

                                           (M
                                                 §,
                                          .8

-------
  o
  in

  en*
        §
  o
  o

   JO
   g 4J f»


   O C '"
   S. O 4)
   jc N x:
   OO *J
       CM
 o
.o
  I

 (V

§
•
en


1 i j— •• — i i
o
o
I
w
S^lOA j_Q| X
515.2-39
O
O
1
*«


-"•• ••• •
o
o
0
0



-------
THIS PAGE LEFT BLANK INTENTIONALLY
              515.2-40

-------
     METHOD 524.2.   MEASUREMENT OF PUR6EABLE ORGANIC COMPOUNDS IN WATER BY
                    CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY
                                 Revision 4.1
                          Edited by J.W.  Munch (1995)
A. AT ford-Stevens, J.W. Eichelberger, W.L. Budde - Method 524, Rev. 1.0 (1983)

R.W. Slater, Jr. -  Revision 2.0 (1986)

J.W. Eichelberger, and W.L. Budde - Revision 3.0 (1989)

J.W. Eichelberger, J.W. Munch, and T.A. Bellar - Revision 4.0 (1992)
                     NATIONAL EXPOSURE  RESEARCH LABORATORY
                      OFFICE OF RESEARCH AND DEVELOPMENT
                     U.S. ENVIRONMENTAL PROTECTION AGENCY
                            CINCINNATI,  OHIO   45268
                                    524.2-1

-------
                                 METHOD 524.2

            MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN WATER BY
            CAPILLARY  COLUMN  GAS CHROMATOGRAPHY/MASS  SPECTROMETRY


1.    SCOPE AND APPLICATION

     1.1  This is a general  purpose method for the identification and simulta-
          neous measurement  of purgeable volatile organic compounds  in surface
          water,  ground water, and  drinking water in any stage of treatment
          (1,2).   The method is applicable to a wide range of organic com-
          pounds, including  the four trihalomethane disinfection by-products,
          that have sufficiently high volatility and low water solubility to
          be removed from water samples with purge and trap procedures.  The
          following compounds  can  be determined by this'method.

                                              Chemical Abstract  Service
              Analvte                              Registry Number

              Acetone*                                 67-64-1
              Acrylonitrile*                          107-13-1
              Ally!  chloride*                          107-05-1
              Benzene                                  71-43-2
              Bromobenzene                            108-86-1
              Bromochlorom'ethane                       74-97-5
              Bromodichloromethane                      75-27-4
              Bromoform                             .   75-25-2
              Bromomethane                             74-83-9
              2-Butanone*                              78-93-3
              n-Butylbenzene                          104-5J-8
              sec-Butyl benzene                        135-98-8
              tert-Butylbenzene                        98-06-6
              Carbon disulfide*                        75-15-0
              Carbon tetrachloride                      56-23-5
              Chloroacetonitrile*                      107-14-2
              Chlorobenzene                            108-90-7
              1-Chlorobutane*                          109-69-3
              Chlorpethane                             75-00-3
              Chloroform                                67-66-3
              Chloromethane                             74-87-3
              2-Chlorotoluene                           95-49-8
              4-Chlorotoluene                          106-43-4
              Dibromochloromethane                     124-48-1
              l,2-Dibromo-3-chloropropane               96-12-8
              1,2-Dibromoethane                        106-93-4
              Dibromomethane                       '     74-95-3
              1,2-Dichlorobenzene                       95-50-1
              1,3-Dichlorobenzene                      541-73-1
              1,4-Dichlorobenzene                      106-46-7
              trans-l,4-Dichloro-2-butene*             110-57-6
              Dichlorodifluoromethane                   75-71-8

                                   524.2-2

-------
   1,1-Dichloroethane            .            75-34-3
   1,2-Dichloroethane                       107-06-2
   1,1-Dichloroethene                        75-35-4
   cis-l,2-Dichloroethene                   156-59-2
   trans-l,2-Dichloroethene                 156-60-5
   1,2-Dichloropropane                       78-87-5
   1,3-Dichloropropane                   ,   142-28-9
   2,2-Dichloropropane                      590-20-7
   1,1-Dichloropropene                      563-58-6
   1,1-Dichloropropanone*                   513-88-2
   cis-l,3-Dichloropropene               10061-01-5
   trans-l,3-Dichloropropene             10061-02-6
   Diethyl ether*                           60-29-7
   Ethyl benzene                             100-41-4
   Ethyl methacrylate*                      97-63-2
  Hexachlorobutadiene                      87-68-3
  Hexachloroethane*                        67 72 1
  2-Hexanone*                             591-78-6
  Isopropylbenzene                         98-82-8
  4-Isopropyltoluene                       99-87-6
  Methacrylonitrile*                      126-98-7
  Methylacrylate*                          96-33-3
.  Methylene  chloride                       75.09-?
  Methyl  iodide*                           74-88-4
  Methylmethacrylate*                       80-62-6
  4-Methyl.-2-pentanone*                   108-10-1
  Methyl-t-butyr ether*                  1634-04-4
  Naphthalene                              91-20-3
  Nitrobenzene*  '.                           98-95-3
  2-Nitropropane*                           7g_46_g
  Pentachloroethane*                       76-01-7
  Propionitrile*   .                        107-12-0
  n-Propylbenzene                          103-65 1
  Styrene                                  100-42-5
  1,1,1,2-Tetrachloroethane                630-20-6
  1,1,2,2-Tetrachloroethane                 79-34-5
 Tetrachloroethene                        127-18-4
 Tetrahydrofuran*                         109-99-9
 Toluene      •                            inn ao ••» •
 1 o i -r •  i i '  •                          lUO-OO-J
 1,2,3-Trichlorobenzene                   87-61-6
 1,2,4-Trichlorobenzene                   120-82-1
 1,1,1-Trichloroethane                    71-55-6
 1,1,2-Trichloroethane                    79-00-5
 Trichloroethene                          79-01-6
 Trichlorofluoromethane                   75-69-4
 1,2,3-Trichloropropane                   96-18-4
 1,2,4-Trimethylbenzene                   95-63-6
 1,3,5-Trimethylbenzene                  108-67-8
 Vinyl  chloride                            75_01  4
 o-Xylene                                 95-47-6
 m-Xy  ene                                108-38-3
 P-Xylene                                106-42-3

 New Compound in  Revision 4.0

                      524.2-3

-------
     1.2  Method detection limits (MDLs) (3)  are compound,  instrument and
          especially matrix dependent and vary from approximately 0.02 to 1.6
          /zg/L.  The applicable concentration range of this method is primari-
          ly column and matrix dependent, and is approximately 0.02 to 200
          /zg/L when a wide-bore thick-film capillary column is used.   Narrow--
          bore thin-film columns may have a capacity which  limits the range to
          about 0.02 to 20 /zg/L.  Volatile water soluble,  polar compounds
          which have relatively low purging efficiencies can be determined
          using this method.  Such compounds  may be more susceptible to matrix
          effects, and the quality of the data may be adversely influenced.

     1.3  Analytes that are not separated chromatographically, but which have
          different mass spectra and noninterfering quantitation ions (Table
          1), can be identified and measured  in the same calibration mixture
          or water sample as long as their concentrations are somewhat similar
          (Sect. 11.6.2).  Analytes that have very similar mass spectra cannot
          be individually identified and measured in the same calibration
          mixture or water sample unless they have different retention times
          (Sect. 11.6.3).  Coeluting compounds with very similar mass spectra,
          typically many structural isomers,  must be reported as an isomeric
          group or pair.  Two of the three isomeric xylenes and two of the
          three dichlorobenzenes are examples of structural isomers that may
          not be resolved on the capillary column, and if not, must be
          reported as isomeric pairs.  The fnore water soluble compounds (> 2%
          solubility) and compounds with boiling points above 200°C are purged
          from the water matrix with lower efficiencies.  These analytes may
          be more susceptible to matrix effects.

2.   SUMMARY OF METHOD

     2.1  Volatile organic compounds and surrogates with low water solubility
          are extracted  (purged) from the sample matrix by bubbling an inert
          gas through the aqueous sample.  Purged sample components are
          trapped in a tube containing suitable sorbent materials.  When
          purging is complete, the sorbent tube is heated and backflushed with
          helium to desorb the trapped sample components into a capillary gas
          chromatography (GC) column interfaced to a mass spectrometer (MS).
          The column is temperature programmed to facilitate the separation of
          the method analytes which are then detected with the MS.  Compounds
          eluting from the GC column are identified by comparing their
          measured mass  spectra and retention times to reference spectra  and
          retention times in a data base.  Reference spectra and retention
          times for analytes are obtained by the measurement of calibration
          standards under the same conditions used for samples.  Analytes  are
          quantitated using procedural standard calibration  (Sect. 3.14). The
          concentration  of each identified component is measured by relating
          the MS response of the quantitation ion produced by that compound to
          the MS response of the quantitation ion produced by a compound  that
          is used as an  internal standard. I Surrogate analytes, whose
          concentrations are known in every sample, are measured with the  same
          internal standard calibration  procedure.
                                    524.2-41

-------
3.   DEFINITIONS

     3.1  INTERNAL STANDARD (IS) — A pure analyte(s) added to a sample,
          extract, or standard solution in known amount(s) and used to measure
          the relative responses of other method analytes and surrogates  that
          are components of the same sample or solution.   The internal
          standard must be an analyte that is not a sample component.

     3.2  SURROGATE ANALYTE (SA) — A pure analyte(s),  which ,is extremely
          unlikely to be found in any sample, and which is added to a  sample
          aliquot in known amount(s) before extraction  or other processing and
          is measured with the same procedures used to  measure other sample
          components.  The purpose of the SA is to monitor method performance
          with each sample.

     3.3  LABORATORY DUPLICATES (LD1 and LD2) — Two aliquots of the same
          sample taken in the laboratory and analyzed separately with
          identical  procedures.  Analyses of LD1 and LD2  indicates precision
          associated with laboratory procedures, but not  with sample
          collection, preservation,  or storage procedures.

     3.4  FIELD DUPLICATES (FD1 and FD2) — Two separate  samples collected at
          the same time and place under identical  circumstances and treated
          exactly the same throughout field and laboratory procedures. Analy-
          ses of FD1 and FD2 give a measure of the precision associated with
          sample collection,  preservation and storage,  as well  as with labora-
          tory procedures.

     3.5  LABORATORY REAGENT BLANK ,(LRB) — An aliquot  of reagent water or
          other blank matrix that is treated exactly as a sample including
          exposure to all  glassware,  equipment,  solvents,  reagents,- internal
          standards,  ana surrogates  that are used  with  other samples.   The LRB
          is used to determine if method analytes  or other interferences  are
          present in the laboratory  environment,  the reagents,  or the  appara-
          tus.                                                  •

     3.6  FIELD REAGENT BLANK (FRB)  -- An aliquot  of reagent water or  other
          blank matrix that is placed in a sample.container  in  the laboratory
          and treated as a  sample in  all  respects,  including shipment  to  the  •
          sampling site,  exposure to  sampling site conditions,  storage,
          preservation,  and all  analytical  procedures.  The  purpose of the FRB
          is to determine  if method  analytes or other interferences are
          present in  the field environment.

     3.7  LABORATORY  PERFORMANCE CHECK SOLUTION  (LPC) —  A solution of one or
          more  compounds (analytes,  surrogates,  internal  standard,  or  other
          test  compounds)  used to evaluate  the performance of the instrument
          system with respect  to a defined  set of  method  criteria.

     3.8  LABORATORY  FORTIFIED BLANK  (LFB)  — An aliquot  of  reagent water  or
          other blank matrix  to  which known  quantities  of  the method analytes
          are added  in  the  laboratory.   The  LFB  is  analyzed  exactly like  a
          sample,  and its purpose is  to  determine  whether  the methodology  is


                                   524.2-5

-------
         in control, and whether the laboratory is capable of making accurate
         and precise measurements.

    3.9  LABORATORY FORTIFIED SAMPLE MATRIX  (LFM) — An aliquot of an
         environmental  sample to which known quantities of the method
         analytes  are  added  in the laboratory.  The LFM is analyzed exactly
         like  a  sample, and  its purpose  is to determine whether the sample
         matrix  contributes  bias to the  analytical results.  The background
         concentrations of the analytes  in the sample matrix must be
         determined in a separate aliquot and the measured values in the LFM
         corrected for background concentrations.

    3.10 STOCK STANDARD SOLUTION  (SSS) —jA  concentrated  solution containing
         one or  more method  analytes prepared in the laboratory using assayed
         reference materials or purchased from a reputable commercial source.

    3.11 PRIMARY DILUTION STANDARD SOLUTION  (PDS) — A  solution of  several
         analytes prepared  in the laboratory from stock standard solutions
         and diluted  as needed to prepare calibration solutions and other
         needed  analyte solutions.

    3.12 CALIBRATION  STANDARD (CAL)  -- A solution prepared from the primary
         dilution standard  solution  or  stock standard solutions and the
         internal  standards  and  surrogate analytes.  The  CAL  solutions are
         used  to calibrate  the  instrument response  with respect to  analyte
         concentration.                 !
                                         [
    3.13 QUALITY CONTROL  SAMPLE  (QCS)  — A  solution of  method analytes of
         known concentrations which  is  used  to  fortify  an aliquot  of  LRB or
         sample  matrix.'  The QCS  is  obtained from a source  external to the
         laboratory and different from the  source of calibration  standards.
          It is used to check laboratory performance with  externally prepared
         test  materials.                 ;

     3.14  PROCEDURAL STANDARD CALIBRATION —   A calibration  method  where
          aqueous calibration standards are  prepared and processed  (e.g.
          purged,extracted,  and/or derivatized)  in exactly the same manner  as
          a sample.  All steps in the process from addition  of sampling
          preservatives through .instrumental  analyses are  included  in  the
          calibration.  Using procedural  standard calibration compensates for
          any  inefficiencies in the processing procedure.

4.   INTERFERENCES

     4.1  During analysis, major contaminant sources are volatile materials in
          the  laboratory and  impurities  in the inert purging gas and in the
          sorbent  trap.  The  use of Teflon tubing, Teflon thread sealants,  or
          flow controllers with rubber components in the purging device should
          be avoided since such materials out-gas organic compounds which will
          be concentrated in  the trap'during the purge  operation.  Analyses of
        •  laboratory reagent  blanks provide  information about the presence of
          contaminants.  When potential  interfering peaks are noted in labora-
          tory reagent blanks, the analyst should change  the  purge gas source


                                    524.2-6

-------
           and  regenerate  the molecular  sieve  purge gas  filter.  Subtracting
           blank  values  from sample  results  is  not permitted.

     4.2   Interfering contamination may occur  when a  sample containing low
           concentrations  of volatile  organic  compounds  is  analyzed immediately
           after  a  sample  containing relatively high concentrations of volatile
           organic  compounds.  A  preventive  technique  is between-sample rinsing
           of the purging  apparatus  and  sample  syringes with two portions of
           reagent  water.  After  analysis of a  sample  containing high
           concentrations  of volatile  organic compounds, one 017 more laboratory
           reagent  blanks  should  be  analyzed to check  for cross-contamination.

     4.3   Special  precautions must  be taken to determine methylene chloride.
           The  analytical  and sample storage area should be isolated from all
           atmospheric sources of methylene  chloride,  otherwise random back-
           ground levels will result.  Since methylene chloride will permeate
           Teflon tubing,  all GC  carrier gas lines and purge gas plumbing
           should be constructed  of  stainless steel or copper tubing. .
           Laboratory worker's clothing  should  be cleaned frequently since
           clothing previously exposed to methylene chloride fumes during
           common liquid/liquid extraction procedures  can contribute to sample.
           contamination.  .

     4.4   Traces of ketones, methylene  chloride, and  some other organic sol-
           vents can be present even in  the  highest purity methanol.  This is
           another  potential source of contamination,  and should be assessed
           before standards are prepared in  the methanol.

5.   SAFETY

     5.1   The  toxicity or carcinogenicity of chemicals used in this method has
           not  been precisely defined; each  chemical  should be treated as a
           potential health hazard, and  exposure to these chemicals should be
           minimized.  Each laboratory is responsible for maintaining awareness
           of OSHA  regulations regarding safe handling of chemicals used in
           this method.  Additional references  to laboratory safety are
           available (4-6) for the information  of the analyst.

     5.2   The  following method analytes have been tentatively classified as
           known or suspected human or mammalian carcinogens:   benzene,  carbon
           tetrachloride,  1,4-dichlorobenzene,   1,2-dichlorethane,  hexachloro-
           butadiene, 1,1,2,2-tetrachloroethane, 1,1,2-trichloroethane,  chloro-
           form, l,2-dibromoethane,tetrachloroethene,  trichloroethene,  and
           vinyl chloride.  Pure  standard materials and stock standard
           solutions of these compounds  should  be handled in a hood.   A
          NIOSH/MESA approved toxic gas respirator should be worn when the
           analyst handles high concentrations of these toxic compounds.

6.   EQUIPMENT AND SUPPLIES (All  specifications are  suggested.   Catalog
     numbers are included for illustration only.)        '

     6.,1  SAMPLE CONTAINERS— 40-mL to 120-mL screw cap vials each equipped
          with a Teflon faced silicone .septum.  Prior to use,  wash vials and
          septa with detergent and rinse with tap and distilled water.   Allow

                                    524.2-7

-------
     the vials and septa to air dry at r^oom temperature, place in a 105°C
     oven for 1 hr, then remove and allow to cool in an area known to be
     free of organics.

6.2  PURGE AND TRAP SYSTEM — The purge and trap system consists of three
     separate pieces of equipment:  purging device,  trap,  and desorber.
     Systems are commercially available from several sources that meet
     all of the following specifications.

     6.2.1  The all glass purging device (Figure 1)  should be designed to
            accept 25-mL samples with a water column at least 5 cm deep.
            A smaller (5-mL) purging device is recommended if the GC/MS
            system has adequate sensitivity to obtain the method detec-
            tion limits required.  Gaseous volumes above the sample must
            be kept to a minimum (< 15 ml) to eliminate dead volume
            effects.   A glass frit should be installed at the base of the
            sample chamber so the purge gas passes through the water
            column as finely divided bubbles with a  diameter of < 3 mm at
            the origin.  Needle spargers may be used, however, the purge
            gas must be introduced at a point about  5 mm from the base of
            the water column.   The use of a moisture control  device is
            recommended to prohibit much of the trapped water vapor from
            entering the GC/MS and event'ually causing instrumental  prob-
            1 ems.

     6.2.2  The trap (Figure 2)  must be at least 25  cm long and have an
            inside diameter of at least 0.105 in. Starting from the
            inlet,  the trap should contain 1.0 cm of methyl  silicone
            coated packing and the following amounts of adsorbents:   1/3
            of 2,6-diphenylene oxide polymer,  1/3 of silica gel,  and 1/3
            of coconut charcoal.   If it is not necessary to determine
            dichlorodifluoromethane,  the charcoal can be eliminated  and
            the polymer increased to fill  2/3  of the trap.   Before  ini-
            tial  use,  the trap should be conditioned overnight at 180°C
            by backflushing with  an inert gas  flow of at least 20 mL/min.
            Vent the  trap effluent to the room,  not  to the analytical
            column.   Prior to  daily use,  the trap should be conditioned
            for 10 min at 180°C with  backflushing.   The  trap may  be
            vented to  the analytical  column  during daily conditioning;
            however,  the column must be run  through  the temperature
            program prior to analysis of samples. The use of alternative
            sorbents  is acceptable provided  the  data acquired  meets  all
            quality control  criteria described in Section  9,  and  provided
            the  purge  and desorption  procedures  specified  in  Section 11
            of the  method are  not  changed.   Specifically,  the  purging
            time,  the  purge gas flow rate,  and the desorption  time may
            not  be  changed.  Since many of the potential alternate
            sorbents may be thermally stable above 180°C,  alternate  traps
            may  be  desorbed and baked  out  at higher  temperatures  than
            those described in Section  11.   If higher temperatures are
            used, the  analyst  should  monitor the  data for  possible
            analyte and/or  trap decomposition.


                              524.2-8

-------
     6.2.3  The use of the methyl silicone coated packing is recommended,
            but not mandatory.  The packing serves a dual purpose of
            protecting the Tenax adsorbant from aerosols, and also of
            insuring that the Tenax is fully enclosed within the heated
            zone of the trap thus eliminating potential cold spots.
            Alternatively, silanized glass wool may be used as a spacer
            at the trap inlet.

     6.2.4, The desorber (Figure 2) must be capable of rapidly heating
            the trap to 180°C either prior to or at the beginning of the
            flow of desorption gas.  The polymer section of the trap
            should not be heated higher than 200°C or the life expectancy
            of the trap will decrease.  Trap failure is characterized by
            a pressure drop in excess of 3 lb/in2 across the trap during
            purging or by poor bromoform sensitivities.  The desorber
            design illustrated in Fig. 2 meets these criteria.

6.3  GAS CHROMATOGRAPHY/MASS SPECTROMETER/DATA SYSTEM (GC/MS/DS)

     6.3.1  The GC must be capable of temperature programming and should
            be equipped with variable-constant differential  flow control-
            lers so that the column flow rate will remain constant
            throughout desorption and temperature program operation.  If
            the column oven is to be cooled to 10°C or lower,  a  subam-
            bient oven controller will likely be required.  If syringe
            injections of 4-bromofluorobenzene (BFB) will be used, a
            split/splitless injection port is required.

     6.3.2  Capillary GC Columns.  Any gas chromatography column that
            meets the performance specifications of this method may be
            used (Sect. 10.2.4.1).   Separations of the calibration mix-
            ture must be equivalent or better than those described in
            this method.   Four useful  columns have been evaluated, and
            observed compound retention times for these columns are
            listed in Table 2.

            6.3.2.1  Column 1 — 60 m x 0.75 mm ID VOCOL (Supelco, Inc.)
                     glass wide-bore capillary with a 1.5 /^m film thick-
                     ness.

                     Column 2 — 30 m x 0'.53 mm ID DB-624 (J&W Scien-
                     tific,  Inc.) fused silica capillary with a 3 p,m film
                     thickness.

                     Column 3 — 30 m x 0.32 mm ID DB-5 (J&W Scientific,
                     Inc.)  fused silica capillary with a 1 /zm film thick-
                     ness.

                     Column 4 — 75 m x 0.53 mm id DB-624 (J&W Scien-
                     tific,  Inc.) fused, silica capillary with a 3 IM film
                     thickness.

     6.3.3  Interfaces between  the  GC  and  MS.   The interface used depends
            on the column  selected  and the gas flow rate.
                              524.2-9

-------
        6.3.3.1   The wide-bore columns  1,  2,  and  4  have  the  capacity
                 to accept the standard gas  flows from the trap
                 during thermal  desorption,  and chromatography can
                 begin with the onset of thermal  desorption.  Depend-
                 ing on the pumping  capacity  of the MS,  an additional
                 interface between the  end of the column  and  the  MS
                 may be required.  An open split  interface (7) or an
                 all-glass jet separator is  an acceptable interface.
                 Any interface can be used if the performance speci-
                 fications described in this  method (Sect. 9  and  10)
                 can be achieved.  The  end of the transfer line after
                 the interface,  or the  end of the analytical  column
                 if no interface is  usedj  should  be placed within a
                 few mm of the MS ion source.

        6.3.3.2   When narrow bore column ;3 is  used,  a cryogenic
                 interface placed just  in  front of  the column inlet
                 is  suggested.   This interface condenses  the  desorbed
                 sample components in a narrow band on an uncoated
                 fused silica  precolumn using  liquid nitrogen cool-
                 ing.   When  all  analytes have  been  desorbed from  the
                 trap,  the interface is rapidly heated to transfer
                 them to the analytical  column. The end of the ana-
                 lytical column  should  be  placed  within a few mm  of
                 the MS ion  source.  A  potential  problem with this
                 interface is  blockage  of  the  interface by frozen
                 water from  the  trap.   This condition will result in
                 a major loss  in sensitivity  and  chromatographic
                 resolution.

6.3.4  The mass  spectrometer  must be capable  of  electron ionization
       at a nominal  electron  energy of 70 eV.  The  spectrometer must
       be capable of scanning from  35  to  260  amu with a complete
       scan cycle time  (including scan  overhead)  of 2 sec or less.
        (Scan cycle  time = Total MS  data acquisition time in seconds
       divided by number  of scans in the  chromatogram.)   The spec-
       trometer  must  produce  a mass spectrum  that  meets all  criteria
       in Table  3 when 25 ng  or less of 4-bromofluorobenzene (BFB)
       is introduced  into the GC.  An  average spectrum across the
       BFB GC peak  may be used to test  instrument  performance.

6.3.5  An interfaced  data system is required  to  acquire, store,
       reduce, and  output mass spectral data.  The computer software
       should have  the capability of processing  stored GC/MS data by
       recognizing  a  GC peak within any given retention  time window,
       comparing the mass spectra from the GC peak with  spectral
       data in a user-created data base, and generating  a list  of
       tentatively  identified compounds with their retention times
       and scan numbers.  The software must  allow  integration of the
       ion abundance of any specific ion between  specified  time  or
       scan number limits.  The software should also allow  calcula-
       tion of response factors as  defined in Sect. 10.2.6  (or
       construction of a linear or second order regression  calibra-
       tion curve), calculation of response  factor statistics (mean
       and standard deviation), and calculation of concentrations of
                         524.2-10

-------
                 analytes using either the calibration curve or the equation
                 in Sect. 12.
     6.4  SYRINGE AND SYRINGE VALVES
          6.4.1  Two 5-mL or 25-mL glass hypodermic syringes with Luer-Lok tip
                 (depending on sample volume used).
          6.4.2  Three 2-way syringe valves with Luer ends.
          6.4.3  Micro syringes - 10, 100 ill.
          6.4.4  Syringes - 0.5, 1.0, and 5-mL, gas tight with shut-off valve.
     6.5  MISCELLANEOUS
          6.5.1  Standard solution storage containers — 15-mL bottles with
                 Teflon lined screw caps.
7.  REAGENTS AND STANDARDS
    7.1   TRAP PACKING MATERIALS
          7.1.1  2,6-Diphenylene oxide polymer, 60/80 mesh, chromatographic
                 grade (Tenax GC or equivalent).
          7.1.2  Methyl silicone packing (optional) — OV-1 (3%) on Chromosorb
                 W, 60/80 mesh, or equivalent.
          7.1.3  Silica gel — 35/60 mesh, Davison, grade 15 or equivalent.
          7.1.4  Coconut charcoal — Prepare from Barnebey Cheney, CA-580-26
                 lot #M-2649 (or equivalent) by crushing through 26 mesh
                 screen.
     7.2  REAGENTS-
          7.2.1  Methanol — Demonstrated to be free of analytes.
          7.2.2  Reagent water — Prepare reagent water by passing tap water
                 through a filter bed containing about 0.5 kg of activated
                 carbon, by using a water purification system, or by boiling
                 distilled water for 15 min followed by a 1-h purge with inert
                 gas while the water temperature is held at 90 C.   Store in
                 clean, narrow-mouth bottles with Teflon lined septa and screw
                 caps.
          7.2.3  Hydrochloric acid (1+1) — Carefully add measured volume of
                 cone. HC1 to equal volume of reagent water.
          7.2.4  Vinyl chloride — Certified mixtures of vinyl chloride in
                 nitrogen and pure vinyl chloride are available from several
                 sources (for example, Matheson, Ideal Gas Products, and Scott
                 Gases).
          7.2.5  Ascorbic acid — ACS reagent grade, granular.
                                   524.2-11

-------
      7.2.6   Sodium thiosulfate —  ACS  reagent  grade,  granular.

7.3   STOCK  STANDARD SOLUTIONS — These solutions  may  be  purchased  as
      certified  solutions  or prepared from  pure standard  materials  using
      the  following procedures.  One of these solutions is  required for
      every  analyte of concern,  every surrogate, and the  internal stan-
      dard.   A useful  working concentration is  about 1-5  mg/mL.

      7.3.1   Place  about 9.8 ml of  methanol  into a 10-mL  ground-glass
             stoppered volumetric flask.  Allow the flask to  stand,
             unstoppered,  for about 10  miii  or until all alcohol-wetted
             surfaces  have dried and weigh  to the  nearest 0.1 mg.

      7.3.2   If  the analyte is a liquid at  room temperature,  use  a  100-/iL
             syringe and immediately add  two or more drops  of reference
             standard  to the flask.  Be sure that  the  reference standard
             falls  directly into the alcohol without contacting the neck
             of  the flask.   If the  analyte  is a gas at room temperature,
             fill a 5-mL valved gas-tight syringe  with the  standard to the
             5.0-mL mark,  lower the needle  to 5 mm above  the methanol
             meniscus,  and slowly inject  the standard  into  the neck area  .
             of  the flask.   The gas will  rapidly dissolve in the  methanol.

      7.3.3   Reweigh,  dilute to volume,  stopper, then  mix by  inverting the
             flask  several  times.   Calculate the concentration in /KJ//ZL
             from the  net  gain in weight.   When compound  purity is  certi-
             fied at 96% or greater, the  weight can be used without cor-
             rection to calculate the concentration of the  stock  standard.

      7.3..4   Store  stock standard solutions  in  15-mL bottles equipped with
             Teflon lined  screw caps.   Methanol  solutions of acryloni-
             trile,  methyl  iodiae,  and  methyl acrylate are  stable for only
             one week  at 4°C.   Methanol  solutions  prepared  from other
             liquid analytes are stable for  at  least 4 weeks when stored
             at  4°C.  Methanol .solutions prepared  from gaseous analytes
             are not stable for more than 1 week when  stored at < 0°C;  at
             room temperature,  they must  be discarded  after 1 day.

7.4   PRIMARY DILUTION STANDARDS -- Use stock standard solutions  to
      prepare primary  dilution  standard solutions  that contain all  the
      analytes of concern  in methanol or  other  suitable solvent.  The
      primary dilution standards should be  prepared at concentrations that
      can  be  easily diluted to  prepare  aqueous  calibration  solutions that
     will bracket  the working  concentration range.  Store  the primary
     dilution standard solutions with  minimal  headspace  and check  fre-
     quently for signs of deterioration  or evaporation,  especially just
      before  preparing calibration  solutions.   Storage times described for
      stock standard solutions  in Sect. 7.3.4 also apply  to primary
     dilution standard solutions.

7.5   FORTIFICATION  SOLUTIONS  FOR INTERNAL  STANDARD AND SURROGATES

     7.5.1  A solution containing  the  internal  standard  and the  surrogate
            compounds is  required  to prepare laboratory  reagent  blanks

                               524.2-12

-------
             (also  used  as  a  laboratory  performance  check  solution),  and
             to  fortify  each  sample.   Prepare  a  fortification  solution
             containing  fluorobenzene  (internal  standard),  1,2-  dichloro-
             benzene-d4  (surrogate), and BFB (surrogate) in methanol at
             concentrations of  5 /tg/mL of  each  (any  appropriate  concentra
             tion is  acceptable).   A 5-fil  aliquot  of this  solution  added
             to  a 25-mL  water sample volume gives  concentrations  of 1 jug/L
             of  each.  A 5-#L aliquot of this  solution  added to  a 5-mL
             water  sample volume gives a concentration  of  5 /jg/L  of each.
             Additional  internal standards and surrogate analytes are
             optional.   Additional  surrogate compounds  should  be  similar
             in  physical  and  chemical characteristics to the analytes of
             concern.

7.6  PREPARATION OF  LABORATORY REAGENT BLANK  (LRB)  —  Fill a  25-mL  (or
     5-mL) syringe with reagent water and adjust  to the mark  (no air
     bubbles).  Inject  an appropriate volume of the fortification  solu-
     tion containing the internal  standard and surrogates through  the
     Luer Lok valve  into the reagent water.   Transfer  the LRB to the
     purging device.  See Sect. 11.1.2.

7.7  PREPARATION OF  LABORATORY FORTIFIED BLANK - Prepare this exactly
     like a calibration standard (Sect.  7.8).   This is a calibration
     standard that is treated as a sample.

7.8  PREPARATION OF CALIBRATION STANDARDS

     7,8.1  The number of calibration  solutions (CALs)  needed depends on
            the calibration range desired.  A minimum of three CAL solu-
            tions is required to calibrate a range of a factor of 20 in
            concentration.   For a factor of  50,  use at least four stan-
            dards,  and for a  factor of 100 at  least five standards.  One
            calibration standard should  contain each analyte of concern
            at a concentration  of 2-10 times  the method detection limit
            (Tables 4,  5,  and 7)  for that  compound.   The other CAL stan-
            dards  should contain  each  analyte  of concern at concentra-
            tions  that  define the  range  of the method.   Every  CAL solu-
            tion contains the internal  standard  and  the surrogate com-
            pounds  at the same  concentration  (5  /ig/L suggested for a 5-mL
            sample;  1 /zg/L  for  a  25-mL  sample).

     7.8.2  To prepare  a .calibration standard,  add an appropriate volume
            of a primary dilution  standard containing all  analytes  of
            concern  to  an  aliquot  of acidified (pH 2) reagent  water in  a
            volumetric  flask.   Also add  an appropriate  volume  of internal
            standard  and surrogate  compound solution from  Sect.  7.5.1.
            Use  a microsyringe  and  rapidly inject  the methanol  solutions
            into the  expanded area  of the  filled volumetric  flask.
            Remove  the  needle as quickly as possible after injection.
            Mix  by  inverting  the flask three times only.   Discard the
            contents  contained  in the neck of  the  flask.   Aqueous stan-
            dards are not stable in a volumetric flask  and  should be
            discarded after 1 hr unless transferred  to  a sample  bottle
           and  sealed  immediately.  Alternately,  aqueous  calibration

                             524.2-13

-------
                 standards may be prepared in a gas tight, 5 mL or 25 mL sy-
                 ringe.  NOTE: If'unacidified samples are being analyzed for
                 THMs only, calibration standards should be prepared without
                 acid.

8.   SAMPLE COLLECTION. PRESERVATION. AND STORAGE

     8.1  SAMPLE COLLECTION AND DECHLORINATION

          8.1.1  Collect all samples in duplicate.  If samples, such as fin-
                 ished drinking water, are suspected to contain residual chlo-
                 rine, add about 25 mg of ascorbic acid per 40 mL of sample to
                 the sample bottle before filling.  If analytes that are gases
                 at room temperature (such as vinyl chloride), or analytes in
                 Table 7 are not to be determined, sodium thiosulfate is
                 recommended to reduce the residual chlorine.  Three milli-
                 grams of sodium thiosulfate should be added for each 40 mL of
                 water sample.
                 NOTE:  If the residual chlorine is likely to be present > 5
                 mg/L, a determination of the amount of the chlorine may be
                 necessary.  Diethyl-p-phenylenediamine (DPD) test kits are
                 commercially available to determine residual chlorine in the
                 field.  Add an additional 25 mg of ascorbic acid or 3 mg of
                 sodium thiosulfate per each 5 mg/L of residual chlorine.

          8.1.2  When sampling from a water tap, open the tap and allow the
                 system to flush until the water temperature has stabilized
                 (usually about 10 min).  Adjust the flow to about 500 mL/min
                 and collect duplicate samples containing the desired dechlo-
                 rinating agent from the flowing stream.

          8.1.3  When sampling from an open body of water, partially fill a
                 1-quart wide-mouth bottle or 1-L beaker with sample from a
                 representative area.  Fill duplicate sample bottles contain-
                 ing the desired dechlorihating agent with sample from the
                 larger container.

          8.1.4  Fill sample bottles to overflowing, but take care not to
                 flush out the rapidly dissolving dechlorinating agent.  No
                 air bubbles should pass through the sample as the bottle is
                 filled, or be trapped in the sample when the bottle is
                 sealed.

     8,2 SAMPLE PRESERVATION

          8.2.1  Adjust the pH of all samples to < 2 at the time of collect-
                 ion, but after dechlorination, by carefully adding two drops
                 of 1:1 HC1 for each 40 mL of sample.  Seal the sample bot-
                 tles, Teflon face down, anci mix for 1 min.  Exceptions to the
                 acidification requirement are detailed in Sections 8.2.2 and
                 8.2.3. NOTE: Do not mix the ascorbic acid or sodium thiosul-
                 fate with the HC1 in the sample bottle prior to sampling.


                                   524.2-14

-------
8.2.2  When sampling for THM analysis only, acidification may be
       omitted if sodium thiosulfate is used to dechlorinate the
       sample.  This exception to acidification does not apply if
       ascorbic acid is used for dechlorination.
8.2.3
8.2.4
                  If a sample foams vigorously when HC1 is added, discard that
                  sample.  Collect a set of duplicate samples but do not acidi-
                  fy them.  These samples must be flagged as "not acidified"
                  and must be stored at 4°C or below.  These samples must be
                  analyzed within 24 hr of collection time if they are to be
                  analyzed for any compounds other than THMs.

                  The samples must be chilled to about 4°C when  collected and
                  maintained at that temperature until analysis.  Field samples
                  that will  not be received at the laboratory on the day of
                  collection must be packaged for shipment with  sufficient ice
                  to ensure  that they will  arrive at the laboratory with a
                  substantial  amount of ice remaining in the cooler.

      8.2   SAMPLE'STORAGE                                                 •    '

           8.2.1   Store samples at < 4°C  until  analysis.   The sample  storage
                  area must  be free of organic solvent vapors and direct or
      .  ".         intense light.

           8.2.2   Analyze all  samples within 14 days of collection.   Samples
                  not analyzed within this  period must be  discarded  and re-
                  placed.                 '

      8.3   FIELD REAGENT  BLANKS (FRB)

           8.3.1   Duplicate  FRBs  must  be  handled  along with  each sample set,
                  which  is composed  of the  samples  collected  from the  same
                  general  sample  site  at  approximately the same  time.   At the
                  laboratory,  fill  field  blank sample  bottles with reagent
                  water and  sample  preservatives,  seal,  and  ship to the sam-
                  pling site along with empty  sample  bottles  and back  to  the
                  laboratory with.filled  sample bottles.   Wherever a  set  of
                  samples  is shipped  and  stored,,  it  is.accompanied by  appropri-
                  ate  blanks.   FRBs must  remain hermetically  sealed until
                  analysis.

          8.3.2   Use  the same  procedures used  for  samples to add  ascorbic  acid
                  and  HC1 to blanks  (Sect. 8.1.1).  The  same batch of  ascorbic
                  acid and HC1  should  be  used  for the  field reagent blanks  as
                  for  the field samples.

9. QUALITY CONTROL                           .

     9.1  Quality control (QC) requirements are the initial demonstration of
          laboratory capability followed by regular analyses of laboratory
          reagent blanks, field reagent blanks, and laboratory  fortified
          blanks.   A MDL for each analyte must also be determined.  Each

                                   524.2-15

-------
     laboratory must maintain records to document the quality of the data
     generated.  Additional quality control practices are recommended.

9.2  Initial demonstration of low system background.  Before any samples
     are analyzed, it must be demonstrated that a laboratory reagent
     blank  (LRB) is reasonably free of contamination that would prevent
     the determination of any analyte of concern.  Sources of background
     contamination are glassware, purge gas, sorbents, reagent water, and
     equipment.  Background contamination must be reduced to an accept-
     able level before proceeding with the next section.  In general,
     background from method analytes should be below the method detection
     limit.

9.3  Initial demonstration of laboratory accuracy and precision.  Analyze
     four to seven replicates of a laboratory fortified blank containing
     each analyte of concern at a concentration in the range of 2-5 fig/L
     depending upon the calibration range of the instrumentation.

     9.3.1  Prepare each replicate by adding an appropriate aliquot of a
            quality control sample to reagent water.  It is recommended
            that a QCS from a source different than the calibration
            standards be used for this set of LFBs, since it will serve
            as a check to verify the accuracy of the standards used to
            generate the calibration curve.  This is particularly useful
            if the laboratory is using the method for the first time, and
            has no historical data base for standards.  Prepare each
            replicate by adding an appropriate aliquot of a quality
            control sample to reagent water.  Also add the appropriate
            amounts of internal standard and surrogates.  If it is ex-
            pected that field samples will contain a dechlorinating agent
            and HC1, then  add these to the LFBs  in the same amounts pro-
            scribed in Sect. 8.1.1.  If only THMs are to be determined
            and field samples do not contain HC1, then do not acidify
            LFBs.  Analyze each replicate  according to the procedures de-
            scribed in Section 11.

     9.3.2  Calculate the measured concentration of each analyte  in each
            replicate, the mean concentration of each analyte in  all
            replicates, and mean accuracy  (as mean percentage of  true
            value) for each analyte, and the precision  (as relative
            standard deviation, RSD) of the measurements for each
            analyte.
                                      i
                                      I
     9.3.3  Some  analytes, particularly early eluting gases and late
            eluting higher molecular weight compounds, will be measured
            with  less accuracy and precision than other analytes.  Howev-
            er, the accuracy and precision  for all  analytes must  fall
            within the limits expressed below.   If  these criteria are not
            met for an analyte of  interest, take remedial action  and
            repeat the measurements  for that analyte  until  satisfactory
            performance is achieved.  For  each analyte, the mean  accuracy
            must  be 80-120%  (i-e-  an accuracy of ±  20%).  The precision

                              524.2-16

-------
             of the recovery (accuracy) for each analyte must be less than
             twenty percent (<20%).   These criteria are different than the
             ± 30/0 response factor criteria specified in Sect. 10.3.5
             The criteria differ, because the measurements in Sect.'gis.S
             as part of the initial  demonstration of capability are meant
             to be more stringent than the continuing calibration measure-
             ments in Sect. 10.3.5.

      9.3.4  To determine the MDL, analyze a minimum of 7 LFBs prepared at
             a low concentration.  MDLs in Table 5 were calculated from
             samples fortified from  0.1-0.5 /jg/L, which can be used as a
             guide,  or use calibration data to estimate a concentration
             for .each analyte that will yield a peak with a 3-5 signal  to
             noise response.   Analyze the 7 replicates  as described in
             Sect.11,  and on  a schedule that results in the analyses being
             conducted over several  days.   Calculate the mean accuracy and
             standard deviation for  each analyte.  Calculate  the  MDL usinq
             the equation in  Sect.  13.

      9.3.5  Develop and  maintain 'a  system of control, charts  to plot the
           •  precision and accuracy  of analyte and  surrogate  measurements
             as  a  function of  time.   Charting  surrogate  recoveries  is  an
             especially valuable  activity  because surrogates  are  present
             in  every  sample  and  the  analytical  results  will  form a  sig-
             nificant  record of data  quality.

9.4   Monitor  the  integrated areas of the  quantitation  ions of the  inter-
      nal standards  and surrogates (Table  1)  in  all  samples,  continuing
      calibration checks,  and blanks.   These  should  remain  reasonably
      constant over  time.   An abrupt  change may  indicate  a matrix effect
      or an instrument  problem. If a  cryogenic  interface  is utilized  it
      may indicate an  inefficient transfer  from  the  trap  to the column.
      These samples must  be reanalyzed  or  a laboratory fortified  duplicate
      sample analyzed  to  test for matrix effect.  A  more  gradual  drift of
      more  than  50%  in  any  area is indicative of  a loss  in sensitivity
      and the problem must  be found and corrected.                    '

9.5   LABORATORY REAGENT  BLANKS (LRB) - With each batch  of samples pro-
      cessed as  a group within a work shift, analyze a LRB to determine
      the background system contamination.

9.6  Assessing  Laboratory Performance.  Use the procedures and criteria
      in Sects.  10.3.4 and 10.3.5 to evaluate the accuracy of the measure-
     ment of the laboratory fortified blank (LFB), which must be analyzed
     with each batch of samples that is processed as a  group within a
     work shift.  If more than 20 samples are in a work shift batch
     analyze  one LFB per 20 samples.   Prepare the LFB with the concentra-
     tion of  each  analyte that was used in the Sect. 9.3.3 analysis.  If
     the acceptable accuracy for this measurement (±30%) is not achieved
     the problem must be  solved before additional samples may be  reliably
     analyzed.  Acceptance criteria  for the IS and surrogate given in
     Sect.10.3.4 also applies  to  this LFB.

                              524.2-17

-------
          Since the calibration check sample in Sect.  10.3.5 and the  LFB are
          made the same way and since procedural  standards are used,  the
          sample analyzed here may also be used as a calibration check in
          Sect. 10.3.5.  Add the results of the LFB analysis to the control
          charts to document data quality. ,

     9.7  If a water sample is contaminated with  an analyte, verify that it  is
          not a sampling error by analyzing a field reagent blank.  The
          results of these analyses will help define contamination  resulting
          from field sampling, storage and transportation activities.   If the
          field reagent blank shows unacceptable  contamination, the analyst
          should identify and eliminate the contamination.

     9.8  At least quarterly, replicate LFB data  should be evaluated  to
          determine the precision of the laboratory measurements.  Add these
          results to the ongoing control charts to document data quality.

     9.9  At least quarterly, analyze a quality control sample (QCS)  from an
          external source.  If measured analyte concentrations are  not of
          acceptable accuracy, check the entire analytical procedure  to locate
          and correct the problem source.

     9.10 Sample matrix effects have not been observed when this method is
          used with distilled water, reagent water, drinking water, or ground
          water.  Therefore, analysis of a laboratory fortified sample matrix
          (LFM) is not required unless the criteria in Section 9.4  are not
          met.  If matrix effects are observed or suspected to be causing low
          recoveries, analyze a laboratory fortified matrix sample  for that
          matrix.  The sample results' should be flagged and the LFM results
          should be reported with them.

     9.11 Numerous other quality control measures are incorporated  into  other
          parts of this procedure, and serve to alert the analyst to  potential
          problems.

10.  CALIBRATION AND STANDARDIZATION

     10.1 Demonstration and documentation of acceptable initial calibration  is
          required before any samples are analyzed.  In addition, acceptable
          performance must be confirmed intermittently throughout analysis of
          samples by performing continuing calibration checks.  These checks
          are required at the beginning of each work shift, but no  less than
          every 12 hours.  Additional periodic calibration checks are good
          laboratory practice. , It is highly recommended that an additional
          calibration check be performed at the end of any cycle of continuous
          instrument operation, so that each set of field samples is  bracketed
          by calibration check standards.  NOTE:   Since this method uses
          procedural standards, the analysis of the laboratory fortified
          blank, which is required in Sect, 9.6,  may be used here as  a cali-
          bration check sample.

     10.2 INITIAL CALIBRATION

                                   524.2-18

-------
 10.2.1 Calibrate the mass and abundance scales of the MS with cali-
        bration compounds and procedures prescribed by the manufac-
        turer with any modifications necessary to meet the require-
        ments in Sect. 10.2.2.

 10.2.2 Introduce into the GC (either by purging a laboratory reagent
        blank or making a syringe injection)  25 ng or less of BFB and
        acquire mass  spectra for m/z 35-260 at 70 eV (nominal).   Use
        the  purging procedure and/or GC  conditions given  in Sect.  11.
        If the spectrum does not meet all  criteria in Table 3,  the MS
    .    must be returned and adjusted to meet all  criteria before
        proceeding with calibration.   An average spectrum across  the
       .GC peak may be used  to evaluate  the performance of the sys-
        tem.

 10.2.3  Purge a medium CAL solution,  (e.g.,  10-20  /ig/L) using the
        procedure given in Sect.  11.

 10.2.4  Performance criteria for calibration  standards.   Examine  the
        stored  GC/MS  data  with the data  system software.   Figures  3
        and  4 shown acceptable total  ion  chromatograms.

        10.2.4.1  GC performance.  Good column  performance  will pro-
                 duce  symmetrical peaks  with  minimum  tailing  for most
                 compounds.   If peaks  are  unusually  broad,  or if
                 there  is  poor resolution  between  peaks, the  wrong
                 column  has .been  selected  or  remedial  action  is
                 probably  necessary  (Sect. 10.3.6) .

        10.2.4.2  MS sensitivity.  The GC/MS/DS peak  identification
                 software  should  be able to recognize  a GC  peak in
                 the appropriate  retention time window for  each of
                 the compounds  in calibration  solution, and make
                 correct tentative identifications.   If fewer than
                 99% of the compounds are  recognized,  system mainte-
                 nance  is  required.   See Sect. 10.3.6.

10.2.5 If all performance criteria are met, purge an aliquot of each
       of the other CAL solutions using the same GC/MS conditions.

10.2.6 Calculate a response factor (RF)  for each analyte  and isqmer
       pair  for each CAL solution using the internal standard fluor-
       obenzene.  Table 1 contains  suggested quantitation ions for
       all compounds.  This calculation is supported in acceptable
       GC/MS data system software (Sect. 6.3.5), and many other
       software programs.  RF is a  unitless number, but units used
       to express quantities of analyte and internal standard must
       be equivalent.

                RF =   (AxHQ.Js)
                         524.2-19

-------
            where:   Ax  = integrated abundance of the quantitation ion
                            of the analyte.
                     Afs = integrated abundance of the quantitation ion
                            of the internal standard.
                     Qx  = quantity of analyte purged in nanograms or
                            concentration units.
                     Qjs = quantity of internal  standard purged  in ng  or
                            concentration units.

            10.2.6.1 For each analyte and surrogate, calculate the mean
                     RF from analyses of CAL solutions.  Calculate the
                     standard deviation (SD) and the relative standard
                     deviation (RSD) from each mean:  RSD = 100 (SD/M).
                     If the RSD of any analyte or surrogate mean  RF
                     exceeds 20%, either analyze additional aliquots of
                     appropriate CAL solutions to obtain an acceptable
                     RSD of RFs over the entire concentration range, or
                     take action to improve GC/MS performance Sect.
                     10.3.6).  Surrogate compounds are present at the
                     same concentration on every sample, calibration
                     standard, and all types of blanks.

     10.2.7 As an alternative to calculating mean response factors and
            applying the RSD test, use the GC/MS data system software or
            other available software to generate a linear or second order
            regression calibration curve,  by plotting A/Ais  vs. Qx.


10.3 CONTINUING CALIBRATION CHECK — Verify the MS tune and initial
     calibration at the beginning of each  12-hr work shift during which
     analyses are performed using the following procedure. Additional
     periodic calibration checks are good  laboratory practice.   It is
     highly recommended that an additional  calibration check be performed
     at the end of any cycle of continuous instrument operation,  so that
     each set of field samples is bracketed by calibration check stan-
     dards.

     10.3.1 Introduce into the GC (either  by purging a laboratory reagent
            blank or making a syringe injection)  25 ng or less  of BFB and
            acquire a mass spectrum that includes data for m/z  35-260.
            If the spectrum does not meet  all  criteria (Table 3),  the MS
            must be returned and adjusted  to meet all criteria  before
            proceeding with the continuing calibration check.

     10.3.2 Purge a CAL solution  and analyze with the same conditions
            used during the initial  calibration.   Selection of  the con-
            centration level  of the calibration check standard  should  be
            varied so that the calibration  is  verified at more  than one
            point over the course of several  days.

     10.3.3 Demonstrate acceptable performance  for the criteria  shown  in
            Sect.  10.2.4.

                              524.2-20

-------
10.3.4 Determine that the absolute areas of the quantitation ions of
       the internal standard and surrogates have not decreased by
       more than 30% from the areas measured in the most recent
       continuing calibration check,  or by more than 50% from the
       areas measured during initial  calibration.  If these are.as
       have decreased by more than these amounts, adjustments must
       be made to restore system sensitivity.  These adjustments may
       require cleaning of the MS ion source, or other maintenance
       as indicated in Sect. 10.3.6,  and recalibration.  Control
       charts are useful aids in documenting system sensitivity
       changes.

10.3.5 Calculate the RF for each analyte of concern and surrogate ,
       compound from the data measured in the continuing calibration
       check.  The RF for each analyte and surrogate must .be within
       30% of the mean value measured in the initial calibration.
       Alternatively, if a linear or second order regression is
       used, the concentration measured using the calibration curve
       must be within 30% of the true value of the concentration in'
       the calibration solution.  If these conditions do not exist,
       remedial action must be taken which may require recalibrati-
       on.  All data from field samples obtained after the last
       successful calibration check standard, should be considered
       suspect.  After remedial action has been taken, duplicate
       samples should be analyzed if they are available.

10.3.6 Some possible remedial actions.  Major maintenance such as
       cleaning an ion source, cleaning quadrupole rods, etc. re-
       quire returning to the initial calibration step.

       10.3.6.1 Check and adjust GC and/or f'S operating conditions;
                check the MS resolution,  and calibrate the mass
                scale.

       10.3.6.2 Clean or replace the splitless injection liner;
                silanize a new injection liner.  This applies only
                if the injection liner is an integral part of the
                system.

       10.3.6.3 Flush the GC column with solvent according to manu-
                facturer's instructions.

       10.3.6.4 Break off a short portion (about 1 meter) of the
                column from the end near the injector; or replace GC
                column.  This action will cause a slight change in
                retention times.  Analyst may need to redefine
                retention windows.

       10.3.6.5 Prepare fresh CAL solutions,  and repeat the initial
                calibration step.                       .

       10.3.6.6 Clean the MS ion source and rods (if a quadrupole).

                         524.2-21

-------
                 10.3.6.7 Replace any components that allow analytes to come
                          into contact with hot metal surfaces.

                 10.3.6.8 Replace the MS electron multiplier, or any other
                          faulty components.;

                 10.3.6.9 Replace the trap, especially when only a few com-
                          pounds fail the criteria in Sect. 10.3.5 while the
                          majority are determined successfully.  Also check
                          for gas leaks in the purge and trap unit as well as
                          the rest of the analytical system.
                                            I
     10.4 Optional calibration for vinyl chloride using a certified gaseous
          mixture of vinyl chloride in nitrogen can be accomplished by the
          following steps.

          10.4.1 Fill the purging device with 25.0 ml (or 5-mL) of reagent
                 water or aqueous calibration standard.

          10.4.2 Start to purge the aqueous mixture.  Inject a known volume  ,
                 (between 100 and 2000 /zL) of the calibration gas (at room
                 temperature) directly into the purging device with a gas
                 tight syringe.  Slowly inject the gaseous sample through a
                 septum seal at the top of the purging device at 2000 /uL/min.
                 If the injection of the standard is made through the aqueous
                 sample inlet port, flush the dead volume with several mL of
                 room air or carrier gas.  Inject the gaseous standard before
                 5 min of the 11-min purge time have elapsed.
          10.4.3 Determine the aqueous equival
                 chloride standard, in
                 where
                          S = 0.102 (C)(V)
                    ent concentration of vinyl
                   injected with the equation:
S = Aqueous equivalent concentration
     of vinyl chloride standard in
C - Concentration of gaseous standard in mg/L (v/v);
V = Volume of standard injected in mL.
11.  PROCEDURE
     11.1 SAMPLE INTRODUCTION AND PURGING
                                            I
          11.1.1 This method is designed for a 25-mL or 5-mL sample volume,
                 but a smaller (5 mL) sample volume is recommended if the
                 GC/MS system has adequate sensitivity to achieve the required
                 method detection limits.  Adjust the helium purge gas flow
                 rate to 40 mL/min.  Attach the trap inlet to the purging
                 device and open the syringe valve on the purging device.

          11.1.2 Remove the plungers from two 25-mL (or 5-mL depending on
                 sample size) syringes and attach a closed syringe valve to
                 each.  Warm the sample to room temperature, open the sample
                                   524.2-22

-------
           .  botMe, and carefully pour the sample into one of the syringe
             barrels to just short of overflowing.  Replace the syringe
             plunger, invert the syringe, and compress the sample   Open
             the syringe valve and vent any residual  air while adjusting
             the sample volume to 25.0-mL (or 5-mL).   To all samples,
             blanks, and calibration standards,  add 5-//L (or an appropri-
             ate volume) of the fortification solution containing the
             internal standard and the surrogates to  the sample through
             the syringe valve.  Close the valve.  Fill  the second syringe
             in an identical manner from the same sample bottle.   Reserve
             this second syringe for a reanalysis if  necessary.

      11.1.3 Attach the sample syringe valve to  the syringe valve on the
             purging device.  Be sure that the trap is cooler than 25°C,
             then open  the sample syringe valve  and inject  the sample into
             the purging chamber.   Close both valves  and initiate purging
             Purge the  sample for 11.0 min at ambient  temperature.

      11.1.4 Standards  and samples  must be analyzed in exactly the same
             manner.  Room temperature must  be reasonably constant,  and
             changes  in  excess  of 10°F will  adversely  affect  the  accuracy
             and precision of the method.

 11.2  SAMPLE DESORPTION

      11.2.1  Non-cryogenic interface  —  After  the 11-min purge, place  the
             purge  and trap  system  in  the  desorb  mode  and preheat  the  trap
             to  180 C without a flow of desorption gas.  Then simultan-
             eously  start  the  flow  of  desorption  gas at  a flow  rate  suit-
             able  for the  column being  used  (optimum desorb flow  rate  is
             15  mL/min)  for  about 4 min, begin the  GC  temperature  program
             and  start data  acquisition.

      11.2.2  Cryogenic interface — After  the  11-min purge,  place  the
             purge and trap  system  in the  desorb  mode, make sure the
             cryogenic interface is a -150°C or lower,  and  rapidly heat
             the trap to 180°C while backflushing with an inert gas at
             4 mL/min for  about 5 min.  At the end of the 5 min desorption
             cycle, rapidly  heat the cryogenic trap to 250°C,  and  simulta-
             neously begin the temperature program of the gas chromato-
             graph, and start data acquisition.

     11.2.3  While the trapped components  are being introduced into the
            gas chromatograph (or cryogenic interface),  empty the purging
            device using the sample syringe and  wash  the chamber with two
             25-mL flushes of reagent water1.   After the purging device has
            been emptied,  leave syringe valve open to allow the purge gas
            to vent through the sample introduction needle.

11.3 GAS CHROMATOGRAPHY/MASS SPECTROMETRY — Acquire  and store data over
     the nominal mass range 35-260 with a total .cycle time  (including
     scan overhead time) of 2 sec or less.  If water,  methanol, or carbon
     dioxide cause a background problem, start at 47  or 48  m/z.  If

                              524.2-23

-------
    ketones are to be determined, data must be acquired starting at m/z
    43. Cycle time must be adjusted to measure five or more spectra
    during the elution of each GC peak,  Suggested temperature programs
    are provided below.  Alternative temperature programs can be used.

    11.3.1 Single ramp linear temperature program for wide bore column 1
           and 2 with a jet  separator.  Adjust the helium carrier gas
           flow rate to within the capacity of the separator, or about
           15 mL/min.  The column temperature is reduced 10°C and held
           for 5 min from the beginning of desorption, then programmed
           to 160°C at 6°C/min,  and held until  all  components  have
           eluted.

    11.3.2 Multi-ramp temperature  program for wide bore column  2 with,,.
           the open split interface.  Adjust the helium carrier :gas  flow
           rate to about 4.6 mL/min.  Jhe column temperature  is  reduced
           to 10°C and held  for 6 min from the beginning of desorption,
           then heated to 70°C at 10°/m,in,  heated to 120°C  at  5°/min,
           heated to  180° at 8°/min,  and held at 180° until  all  com-
           pounds have eluted.

    11.3.3 Single ramp linear  temperature program  for  narrow  bore  column
           3 with a cryogenic  interface.  Adjust the helium carrier  gas
           flow rate  to  about  4  mL/min.  The  column  temperature  is
           reduced to  10°C  and  held  for 5 min  from the beginning  of
           vaporization  from the cryogenic  trap, programmed at  6°/min
           for  10 min, then  15°/min  for 5 min  to  145°C,  and held  until
           all  components  have  eluted.

     11 3  4*Multi-ramp temperature program  for wide bore column  4 with
           the  open  split  interface.   Adjust  the  helium carrier gas  flow
            rate  to  about 7.0 mL/min.   The  column  temperature  is - 10°C
            and  held  for  6  min.  from beginning of desorption,  then heated
            to  100°C  at 10°C/min, heated to  200°C  at 5°C/min  and held at
            200°C  for 8 min  or until  all compounds  of interest had elut-
            ed.                       .  !

11.4 TRAP RECONDITIONING — After describing the sample for 4 min,  recon-
     dition the trap by returning the purge and trap system to the
     purge mode.   Wait 15 sec, then close the syringe valve on the
     purging  device to begin gas flow through the trap.  Maintain the
     trap temperature at 180°C.   Maintain the  moisture control  module, if
     utilized,  at 90°C to remove residual water.  After approximately 7
     min,  turn off the trap  heater and open the syringe valve to stop the
     gas flow through the trap.  When the trap is cool, the next sample
     can be analyzed.

11 5 TERMINATION OF DATA ACQUISITION — When all the sample components
     have eluted from the GC, terminate MS data acquisition.  Use appro-
     priate data output software to display full range mass spectra  and
     appropriate plots of ion abundance as a function  of time.   If any
     ion abundance exceeds the system working range, dilute the sample
                               524.2-24

-------
     aliquot in the second syringe with reagent water and analyze the
     diluted aliquot.

11.6 IDENTIFICATION OF ANALYTES — Identify a sample component by
     comparison of its mass,spectrum (after background subtraction)  to a
     reference spectrum in the user-created data base.  The GC retention
  •   time of the sample component should be within three standard
     deviations of the mean retention time of the compound in the
     calibration mixture.

     11.6.1  In general, all ions that are present above 10% relative
            abundance in the  mass spectrum of the standard should be
            present in the mass spectrum of the sample component and
            should agree within absolute 20%.   For example,  if an ion has
            a relative abundance of 30% in the standard spectrum,  its
            abundance in the  sample spectrum should be in the range  of 10
            to 50%.   Some  ions,  particularly the molecular ion,  are  of
            special  importance,  and should be  evaluated even  if they are
            below 10% relative abundance.

     11.6.2  Identification requires expert judgment when sample  compo-
            nerits are not  resolved chromatographically and produce mass
            spectra  containing ions contributed by more than  one analyte.
            When GC  peaks  obviously represent  more than one  sample compo-
            nent (i.e.,  broadened peak  with shoulder(s)  or valley between
            two or more maxima),  appropriate analyte spectra  and back-
            ground spectra can be selected by  examining plots  of charac-
            teristic  ions  for  tentatively  identified components.  When
            analytes  coelute  (i.e.,  only one GC peak is  apparent), the
            identification criteria can  be met  but each  analyte  spectrum
            will  contain extraneous ions  contributed by the coeluting
            compound.   Because purgeable  organic  compounds are relatively
            small  molecules and  produce  comparatively  simple mass spec-
            tra,  this  is not a significant problem for  most method
            analytes.

     11.6.3  Structural  isomers that  produce  very  similar mass spectra  can
            be  explicitly  identified only  if they  have  sufficiently
            different  GC retention  times.  Acceptable resolution is
            achieved  if the height  of the  valley  between two peaks is
            less  than  25%  of the  average height of the two peaks. Other-
            wise,  structural isomers are identified  as  isomeric pairs.
            Two  of the three isomeric xylenes and  two of the three di-
            chlorobenzenes are examples of structural isomers that may
            not  be resolved on the capillary columns.   If unresolved,
            these groups of isomers must be  reported as isomeric pairs.

    11.6.4 Methylene chloride, acetone, carbon disulfide, and other
            background components appear in variable quantities in labo-
           ratory and field reagent blanks, and generally cannot be
           accurately measured.  Subtraction of the concentration in the
           blank from the concentration in the sample is not acceptable


                             524.2-25

-------
                 because the concentration of the background in the blank is
                 highly variable.

12. DATA ANALYSIS AND CALCULATIONS

     12.1 Complete chromatographic resolution is not necessary for accurate
          and precise measurements of analyte concentrations if unique ions
          with adequate intensities are available for quantitation.  Lf the
          response for any analyte exceeds the linear range of the .calibration
          established in Section 10, obtain and dilute a duplicate a duplicate
          sample.  Do not extrapolate beyond the calibration range.

          12.1.1 Calculate analyte and surrogate concentrations, using the
                 multi-point calibration established in Section 10.  Do not
                 use the daily calibration verification data to quantitate
                 analytes in samples.
                          C. =
                                 (Ax)(Qis) 1000
                           X
                                   (Ais)  RF  V
                 where:'   Cx  = concentration  of analyte or surrogate in /ig/L
                                 in the water sample.
                          Ax  = integrated abundance of the quantitation ion
                                 of the analyte in the sample.
                          Ais  = integrated abundance  of the quantitation ion
                                 of the internal standard  in the sample.
                          Qis  = total  quantity (in  micrograms), of internal
                                 standard added to the water sample.
                          V   = original water sample volume in mL.
                          RF  = mean response factor of analyte from the
                                 initial calibration.

          12.1.2 Alternatively, use the GC/MS system software or other
                 available proven software to compute the  concentrations of
                 the analytes and surrogates  from the linear or second order
                 regression curve established in Section 10.  Do not use the
                 daily calibration verification data to quantitate analytes in
                 samples.        .                       .              '

          12.1.3 Calculations should utilize  all available digits of precis-
                 ion, but final reported concentrations should be rounded to
                 an appropriate number of significant figures (one digit of
                 uncertainty).  Experience indicates that  three significant
                 figures may be used for concentrations above 99 /ig/L, two
                 significant figures for concentrations between 1- 99 /ig/L,
                 and one significant figure for lower concentrations.

          12.1.4 Calculate the total trihalomethane concentration by summing
                 the four individual trihalomethane concentrations.

13.  METHOD PERFORMANCE

     13.1 Single laboratory accuracy and precision data were obtained for the
          method analytes using laboratory fortified blanks with analytes at

                                   524.2-26

-------
           concentrations between 0.1 and 5 pg/L.  Results were obtained using
           the four columns specified (Sect. 6.3.2.1) and the open split or jet
           separator (Sect. 6.3.3.1), or the cryogenic interface (Sect.
           6.3.3.2).   These data are shown in Tables 4-8.

      13.2  With these data, method detection limits were calculated using the
           formula  (3):

           MDL = S  t(n.1f1_alpha = Oi?9)

           where:

           V-i.i-aipha • o.99) * studen.V.s t value for the 99% confidence
                                    level  with  n-1  degrees of freedom,

           n = number of  replicates

           S = the  standard deviation of  the
               replicate  analyses.

14.  POLLUTION  PREVENTION

     14.1  No  solvents are  utilized  in this  method  except the  extremely small
           volumes  of methanol needed  to make calibration standards.  The only
           other chemicals  used  in this method  are  the neat materials in
           preparing  standards and sample  preservatives.   All  are used  in •
           extremely  small  amounts and pose  no  threat to  the environment.

15.  WASTE MANAGEMENT

     15.1 There are  no waste management issues involved  with  this method.  Due
          to the nature of this method, the discarded samples are chemically
          less contaminated than when they were collected.

16. REFERENCES

     1.    J.W. Munch, J.W. Eichelberger,  "Evaluation of  48 Compounds for
          Possible Inclusion in USEPA Method 524.2, Revision 3.0: Expansion of
          the  Method Analyte List to a Total of 83 Compounds", J. Chro. Scl
          ,30, 471,1992.                                        	'-

     2.    C. Madding, "Volatile Organic Compounds in Water by Purge and Trap
          Capillary Column GC/MS," Proceedings of the Water Quality Technology
          Conference, American Water Works Association,  Denver,  CO, December
          1984.

     3.    J.A. Glaser,  D.L. Foerst,  G.D.  McKee, S.A. Quave,  and W.L.  Budde,
          "Trace Analyses for Wastewaters",  Environ. Scl. Technol.. 15, 1426,
          1981.

     4.    "Carcinogens-Working with  Carcinogens," Department  of Health
          Education,  and  Welfare,  Public  Health Service,  Center for Disease
          Control,  National Institute for Occupational  Safety and Health,
          Publication No.  77-206,  August  1977.

                                   524.2-27

-------
5.   "OSHA Safety and Health Standards, General Industry," (29CFR1910),
     Occupational Safety and Health Administration, OSHA 2206, (Revised,
     January 1976).

6.   "Safety in Academic Chemistry Laboratories," American Chemical
     Society  Publication, Committee on Chemical Safety, 3rd Edition,
     1979.
                                        i
7.   R.F. Arrendale, R.F. Severson, and O.T. Chortyk, "Open Split Inter-
     face for Capillary Gas Chromatography/Mass Spectrometry," Anal.
     Chem. 1984, 56, 1533.

8.   J.J. Flesch, P.S. Fair, "The Analysis of Cyanogen Chloride in Drink-
     ing Water," Proceedings of Water Quality Technology Conference,
     American Water Works Association, St. Louis, MO., November 14-16,
     1988.
                              524.2-28

-------
 17.  TABLES. DIAGRAMS. FLOWCHARTS. AND VALIDATION DATA


      TABLE  1.   MOLECULAR  WEIGHTS  AND  QUANTITATION  IONS  FOR METHOD ANALYTES
                                          Primary           Secondary
                                       Quantitation       Quantitation
 Compound	jnf	Ion   	ions


      Internal  standard

 Fluorobenzene                 96             96                      77

      Surrogates

 4-Bromofluorobenzene         174            95                 174,176
 l,2-Dichlo.robenzene-d4       150           152                 115,'l50

      Target Analytes

 Acetone                        58            43                      58
 Acrylonitrile                  53            52                      53
 Ally!  chloride                 76   .         76                      49
 Benzene                        78            78                      77
 Bromobenzene                 156           156                 77  153
 Bromochloromethane            128           128                 49'130
 Bromodichloromethane         162            83                 85'l27
 Bromoform                     250           173                 17s'252
 Bromomethane                  94            94                   '  gg
 2-Butanone                     72            43                  57  72
 n-Butylbenzene                134            91                     134
 sec-Butyl benzene              134           105                     134
 tert-Butylbenzene             134           119                      91
 Carbon disulfide               76            76
 Carbon tetrachloride   .       152           117                     119
 Chloroacetonitrile             75            48                      75
 Chlorobenzene                 112           112                 77  114
 1-Chlorobutane       .          92             56                   '   49
 Chloroethane                   64             64                      66
 Chloroform                    118            83                     85
 Chloromethane                  50             50                      52
 2-Chlorotoluene               126            91                    126
 4-Chlorotoluene               126            91                    126
 Dibromochloromethane          206           129                    127
 l,2-Dibromo-3-Chloropropane   234            75                155,157
 1,2-Dibromoethane             186           107                109'l88
Dibromomethane                 172            93                 95^174
 1,2-Dichlorobenzene           146           146                Ill'l48
 1,3-Dichlorobenzene           146           146                111^148
 1,4-Dichlorobenzene           146  .         146                111^148


                                   524.2-29

-------
                             TABLE 1.   (continued)
Compound
MW
   Primary
Quantitation
     Ion
  Secondary
Quantitation
     Ions
trans-l,4-Dich1oro-2-butene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
ci s-1 , 2-Di chl oroethene
trans-1 , 2-Di chl oroethene
1 , 2-Di chl oropropane
1,3-Dichloropropane
2, 2-Dichl oropropane
1 , 1-Di chl oropropene
1,1-Dichloropropanone
ci s-1 ,3-di chl oropropene
trans-1 , 3-di chl oropropene
Di ethyl ether
Ethyl benzene
Ethyl methacryl ate
Hexachl orobutadi ene
Hexachloroethane
2-Hexanone
Isopropyl benzene
4-Isopropyl to! uene
Methacryl onitrile
Methyl acrylate
Methyl ene chloride
Methyl iodide
Methylmethacryl ate
4-Methyl -2-pentanone
Methyl -t-butyl ether
Naphthalene
Nitrobenzene
2-Nitropropane
Pentachloroethane
Propionitrile
n-Propyl benzene
Styrene
1,1,1, 2-Tetrachl oroethane
1,1,2, 2-Tetrachl oroethane
Tetrachl oroethene
Tetrahydrofuran
Toluene
1,2,3-Trichlorobenzene
1,2, 4-Tri chl orobenzene
1,1, 1-Tri chl oroethane
1,1,2-Tri chl oroethane
124
120
98
98
96
96
96
112
112
112
110
126
110
110
74
106
114
258
234
100
120
134
67
86
84
142
100
100
88
128
123
89
200
55
120
104
166
166
164
72
92
180
180
132
132
53
85
63
62
96
96
96
63
76
77
75
43
75
75
59
91
69
225
;117
' 43
;105
119
67
55
. 84
• 142
69
43
73
128
51
46
117
54
91
104
131
83
166
71
92
180
180
97
: 83
88,75
87
65,83
98
61,63
61,98
61,98
112
78
97
110,77
83
110
110
45,73
106
99
260
119,201
58
120
134,91
52
85
86,49
127
99
58,85
57
—
77
—
119,167
—
120
78
133,119
131,85
168,129
72,42
91
182
182
99,61
97,85
                                    524.2-30

-------
                             TABLE 1.  (continued)
Compound
MWa
    Primary
 Quantitation
	Ion
   Secondary
 Quantitation
	Ions
Trichloroethene
Tri chl orof 1 uoromethane
1,2,3-Trichloropropane
1, 2, 4-Trimethyl benzene
1, 3, 5-Trimethyl benzene
Vinyl Chloride
o-Xylene
m-Xylene
p-Xylene
130
136
146
120
120
62
106
106
106
95
101
75
105
105
62
106
106
106:
130,132
.103
77
120
' 120
64
91
91 ,
91
aMonoisotopic molecular weight calculated from the atomic masses of the
 isotopes with the smallest masses.
                                   524.2-31

-------
             TABLE 2.  CHROMATOGRAPHIC RETENTION TIMES FOR METHOD ANALYTES
                   ON THREE COLUMNS WITH FOUR SETS OF CONDITIONS3
Compound
        Retention
Col. lb    Col.  2b
 Time
Col. 2C
(min:sec)
 Col.  3d
                                                                            Col. 4e
     Internal standard

Fluorobenzene                    8:49       6:|27

     Surrogates
4-Bromof1uorobenzene
1,2-Di chlorobenzene-d4

     Target Analvtes
 18:38      15:43
 22:16      19:,08
8:14
18:57
6:44
10:35
17:56
2:01
22:13
20:47
20:17
5:40
15:'52
4:23
8:29
14:53
0:58
I
19:29
18:05
17:34
Acetone
Acrylonitrile
Ally! chloride
Benzene
Bromobenzene
Bromochloromethane
Bromodi chloromethane
Bromoform
Bromomethane
2-Butanone
n-Butylbenzene
sec-Butyl benzene
tert-Butylbenzene
Carbon Disulfide                              :
Carbon Tetrachloride             7:37       5:16
Chloroacetonitrile
Chlorobenzene                    15:46      13:01
1-Chlorobutane
Chloroethane                     2:05       1:01
Chloroform                       6:24       4:48
Chloromethane                    1:38       0:44
2-Chlorotoluene                  19:20      16:25
4-Chlorotoluene                  19:30      16:43
Cyanogen chloride  (8)
Dibromochloromethane             14:23      11:51
l,2-Dibromo-3-Chloropropane      24:32      21:05
1,2-Dibromoethane                14:44      11:50
Dibromomethane                   10:39   •    7:56
1,2-Dichlorobenzene              22:31      19:10
1,3-Dichlorobenzene              21:13      18:08
1,4-Dichlorobenzene              21:33      18:23
t-1,4-Dichloro-2-butene
Dichlorodifluoromethane          1:33       0:42
1,1-Dichloroethane               4:51       2:56
                        14:06
  23:38
  27:25
              8:03
             22:00
             31:21
             35:51


13:30
24:00
12:22
15:48
22:46
4:48

27:32
26:08
25:36

13:10 .

20:40


12:36
3:24
24:32
24:46

19:12

19:24
15:26
27:26
26:22
26:36

3:08
10:48


7:25
16:25
5:38
9:20
15:42
1:17

17:57
17:28
17:19

7:25

14:20

1:27
5:33
0:58
16:44
16:49
1:03
12:48
18:02
13:36
9:05
17:47
17:28
17r38

0:53
4:02
16:14
17:49
16t58
21:32
31:52
20:20
, 23:36
30:32
12:26
19:41
35:41
34:04
33:26
16:30
21:11
23:51
28:26
21:00

20:27
9:11
32:21
32:38

26:57
38:20
27:19
23:22
35:55
34:31
34:45
31:44
7:16
18:46
                                        524.2-32

-------
TABLE 2.  (continued)
Compound
1,2-Dichloroethane
1,1-Dichloroethene
cis-l,2-Dichlproethene
trans- 1 , 2-Di chl oroethene
1 , 2-Di chl oropropane
1,3-Di chl oropropane
2 , 2-Di chl oropropane
1 , 1-Di chl oropropanone
1,1-Dichloropropene
cis-l,3-dichloroprbpene
trans-1 , 3-di chl oropropene
Diethyl ether
Ethyl benzene
Ethyl Methacrylate
Hexachlorobutadiene
Hexachloroethane
Hexanone
Isopropyl benzene
4-Isopropyltol uene
Methacrylonitrile
Methyl acryl ate
Methylene Chloride
Methyl Iodide
Methyl methacryl ate
4-Methyl-2-pentanone
Methyl -t-butyl ether
Naphthalene
Nitrobenzene
2-Nitropropane
Pentachl oroethane
Propionitrile
n-Propyl benzene
Styrene
1,1,1, 2-Tetrachl oroethane
1 , 1 , 2 , 2-Tetrachl oroethane
Tetrachl oroethene
Tetrahydrofuran
Toluene
1,2, 3-Tri chl orobenzene
1,2, 4-Tri chl orobenzene
1,1, 1-Tri chl oroethane
1,1, 2-Tri chl oroethane
Trichloroethene
Tri chl orof 1 uoromethane
1,2, 3-Tri chl oropropane
1,2, 4-Tri methyl benzene
Retention
Col. lb Col. 2b
8:24
2:53
6:11
3:59
10:05
14:02
6:01

7:49
11.58
13.46

15:59

26:59


18:04
21:12


3:36




27:10




19:04
17:19
15:56
18:43
13:44

12:26
27:47
26:33
7:16
13:25
9:35
2:16
19:01
20:20
5:50
1:34
3:54
2:22
7:40
11:19
3:48

5:17



13:23

23:41


15:28
18:31


2:04




23:31




16:25
14:36
13:20
16:21
11:09

10:00
24:11
23:05
4:50
11:03
7:16
1:11
16:14
17:42
Time
Col . 2°
13:38
7:50
11:56
9:54
15:12
18:42
11:52

13:06
16:42
17:54

21:00

32:04


23:18 .
26:30


9:16




32:12




24:20
22:24
20:52
24:04
18:36

17:24
32:58
31:30
12:50
18:18
14:48
6:12
24:08
31:30
(min:sec}
Col. 3a
7:00
2:20
5:04
3:32
8:56
12:29
5:19

7:10



14:44

19:14


16:25
17:38


2:40




19:04




16:49
15:47
14:44
15:47
13:12

11:31
19:14
18:50
6:46
11:59
9:01
1:46
16:16
17:19
Col. 4e
21:31
16:01
19:53
17:54
23:08
26:23
19:54
24:52
21:08
24:24
25:33
15:31
28:37
25:35
42:03
36:45
26:23
30:52
34:27
20:15
20:02
17:18
16:21
23:08
24:38
17:56
42:29
39:02
23:58
33:33
19:58
32:00
29:57
28:35
31:35
26:27
20:26
25:13
43:31
41:26
20:51
25:59
22:42
14:18
31:47
33:33
      524-.2-33

-------
                                 TABLE 2.  (continued)
Compound
       Retention
Col. lb    Col.  2b
 Time      (minisec)
Col. 2C     Col.  3d
Col. 4e
1,3, 5-Tri methyl benzene
Vinyl chloride
o-Xyl ene
m-Xyl ene
p-Xyl ene
19:28
1:43
17:07
16:10
16:07
•16:54 .
0:47
14:31
13:41
•13:41
24:50
3:56
22:16
21:22
21:18
16:59
1:02
15:47
15:18
15:18
32:26
10:22
29:56
28:53
28:53
"Columns 1-4 are those given in Sect. 6.3.2.1; retention times were measured
 from the beginning of thermal desorption from the trap  (columns 1-2, and 4) or
 from the beginning of thermal release from the cryogenic interface  (column 3).

bGC conditions given in Sect. 11.3.1.

CGC conditions given in Sect. 11.3.2.

dGC conditions given in Sect. 11.3.3.

CGC conditions given in Sect. 11.3.4.
                                        524.2-34

-------
                                                                                    I
TABLE 3. ION ABUNDANCE CRITERIA FOR 4-BROMOFLUOROBENZENE (BFB)
        Mass
                          Relative Abundance Criteria
      50
      75
      95
      96
     173
     174
     175
     176
     177
15 to 40% of mass 95
30 to 80% of mass 95
Base Peak, 100% Relative Abundance
5 to 9% of mass 95
< 2% of mass 174
> 50% of mass 95
5 to 9% of mass 174
> 95% but < 101% of mass 174
5 to 9% of mass 176
                              524.2-35

-------
TABLE 4.  ACCURACY AND PRECISION DATA FROM 16-31 DETERMINATIONS OF THE METHOD
          ANALYTES IN REAGENT WATER USING WIDE BORE CAPILLARY COLUMN la

Compound
Benzene
Bromobenzene
Bromochl oromethane
Bromodi chl oromethane
Bromoform
Brqmomethane
n-Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chloroethane
Chloroform
Chl oromethane
2-Chlorotoluene
4-Chlorotoluene
Dibromochl oromethane
1 , 2-Dibromo-3-chl oropropane
1,2-Dibromoethane
Dibromomethane
1 , 2-Di chl orobenzene
1,3-Di chlorobenzene
1 , 4-Di chl orobenzene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1, 2-Di chloroethane
1,1-Dichloroethene
cis-1,2 Dichloroethene
trans-1 , 2-Di chl oroethene
1 , 2-Di chl oropropane
1 , 3-Di chl oropropane
2 , 2-Di chl oropropane
1,1-Dichloropropene
ci s-1 , 2-Di chl oropropene
trans-1 , 2-Di chl oropropene
Ethyl benzene
Hexachlorobutadiene
Isopropyl benzene
4-Isopropyl tol uene
Methyl ene chloride
Naphthalene
n-Propyl benzene
Styrene
True
Cone. A
Range (%
jjja/n
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0


0
0
0
0
0
0
0
0
.1-10
.1-10
.5-10
.1-10
.5-10
.5-10
.5-10
.5-10
.5-10
.5-10
.1-10
.5-10
.5-10
.5-10
.1-10
.1-10
.1-10
.5-10
.5-10
.5-10
.1-10
.5-10
.2-20
.5-10
.5-10
.1-10
.1-10
.5-10
.1-10
.1-10
.1-10
.5-10
.5-10


.1-10
.5-10
.5-10
.1-10
.1-10
.1-100
.1-10
.1-100

Mean
ccuracy
of True
Value)
97
100
90
95
101
95
100
100
102
84
98
89
90
93
90
99
92
83
102
100
93
99
103
. 90
96
95
94
101
93
97
96
86
98


99
100
101
99
95
104
100
102
Rel.
Std.
Dev.
m
5.
5.
6.
6.
6.
8.
7.
7.
7.
8.
5.
9.
6.
8.
6.
8.
7.
19.
3.
5.
6.
, 6.
6.
7.
5.
5.
6.
6.
5.
6.
6.
16.
8.


8.
6.
7.
6.
5.
8.
5.
7.
7
5
4
1
3
2
6
6
3
8
9
0
1
9
2
3
0
9
9
6
2
9,
4
7
3
4
7
7
6
1
0
9
9


6
8
6
7
3
2
8
2
Method
Det.
Limitb
(ua/L)
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.


0.
0.
0.
0.
0.
0.
0.
0.
04
03
04
08
12
11
11
13
14
21
04
10
03
13
04
06
05
26
06
24
03
12
03
10
04
06
12
12
06
04
04
35
10


06
11
15
12
03
04
04
04
                                    524.2-36

-------
                             TABLE 4.  (Continued)
Compound
1,1,1, 2-Tetrachl oroethane
1 , 1 , 2, 2-Tetrachl oroethane
Tetrachl oroethene
Tol uene
1,, 2 ,3-Tri chl orobenzene
1,2,4-Trichlorobenzene
1 , 1 , 1-Tri chl oroethane
1,1, 2-Tri chl oroethane
Trichloroethene
Tri chl orof 1 uoromethane
1,2,3-Trichloropropane
1, 2, 4-Tri methyl benzene
1, 3, 5-Tri methyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
True
Cone.
Range
(UQ/L)
0.5-10
0.1-10
0.5-10
0.5-10
0.5-10
0.5-10
0.5-10
0.5-10
0.5-10
0.5-10
0.5-10
0.5-10
0.5-10
0.5-10
0.1-31
0.1-10
0.5-10
Mean
Accuracy
(% of True
Value)
90
91
89
102
109
108
98
104
. 90
89
108
99
92
98
103
97
104
. Rel.
Std.
Dev.
m
6.8
6.3
6.8
8.0
8.6
8.3
8.1
7.3
7.3
8.1
14.4
8.1
7.4
6.7
7.2
6.5
7.7
Method
Det.
Limit",
(•im/L)
0.05
0.04
0.14
0.11 '
0.03
0.04
0.08 '
0.10
0.19
0.08
0.32
0.13
0.05
0.17
0.1 1/.
0.05 '
0.13
aData obtained by using column 1 with a jet separator interface and',a
 quadrupole mass spectrometer (Sect. 11.3.1) with analytes divided among
 three solutions.

bReplicate samples at the lowest concentration listed in column 2 of this
 table were analyzed.  These results were used to calculate MDLs.
                                   524.2-37

-------
        TABLE 6.   ACCURACY AND PRECISION DA^A FROM  SEVEN  DETERMINATIONS
                      OF THE METHOD ANALYTES IN REAGENT WATER USING WIDE BORE
                      CAPILLARY COLUMN 2a
Compound
Mean Accuracy ,
(% of True
Value, RSD
No.b 2 LLQ/L Cone.) (%)
Mean Accuracy
(% of True
Value,
0.2 itq/L Cone.)
RSD
Internal Standard
Fluorobenzene

Surrogates

4-Bromof1uorobenzene
1,2-Di chlorobenzene-d4

Target Analvtes

Benzene
Bromobenzene
Bromochloromethane
Bromodichloromethane
Bromoform
Bromomethane
n-Butylbenzene
sec-Butyl benzene
tert-Butylbenzene
Carbon tetrachloride
Chlorobenzene
Chloroethane0
Chloroform
Chloromethane
2-Chlorotoluene
4-Chlorotoluene
Di bromochloromethane
1,2-Di bromo-3-chloropropanec
l,2-Dibromoethanec
Dibromomethane
1,2-Di chlorobenzene
1,3-Di chlorobenzene
1,4-Dichlorobenzene
Di chlorodi f1uoromethane
1,1-Di chloroethane
1,2-Di chloroethane
1,1-Dichloroethene
cis-1,2-Dichloroethene
trans-1,2-Di chloroethene
2
3
98
97
37
38
4
5
6
7
39
40
41
8
42
9
10
43
44
11
97
102
99
96
89
55
89
102
101
84
104
97
110
91
89
95
1.8
3.2
                    4.4
                    3.0
                    5.2
                    1.8
                    2.4
                   27.
                    4.8
                    3.5
                    4.5
                    3.2
                    3.1

                    2.0
                    5.0
                    2.4
                    2.0
                    2.7
13
45
46
47
14
15
16
17
18
19
99 2.1
93 2.7
100
4.0
98 4.1
38
25.
97 2.3
102 ! 3.8
90
100
2.2
3.4
92 2.1
96
95
                       113
                       101
                       102
                       100
                        90
                        52
                        87
                       100
                       100
                        92
                       103

                        95
                        d
                       108
                       108
                       100
                                95
                                94
                                87
                                94
                                d
                                85
                               100
                                87
                                89
                                85
1.3
1.7
                        1.8
                        1..?
                        2.9
                        1.8
                        2.2
                        6.7
                        2.3
                        2.8
                        2.9
                        2.6
                        1.6

                        2.1

                        3.1
                        4.4
                        3.0
                                    2.2
                                    5.1
                                    2.3
                                    2.8

                                    3.6
                                    2.1
                                    3.8
                                    2.9
                                    2.3
                                   524.2-40

-------
                             TABLE 6.  (Continued)
Compound
1,2-Dichloropropane
1,3-Dichloropropane
2,2-Dichloropropanec
l,l-Dichloropropenec
cis-l,3-Dichloropropenec
trans-l,3-Dichloropropene
Ethyl benzene
Hexachl orobutadi ene
Isopropyl benzene
4-Isopropyltoluene
Methylene chloride
Naphthalene
n-Propyl benzene
Styrene
1,1,1, 2-Tetrachl oroethane
1,1,2, 2-Tetrachl oroethane
Tetrachl oroethene
Toluene
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1 , 1 , 1-Tri chl oroethane
1,1,2-Trichl oroethane
Trichloroethene
Tri chl orof 1 uoromethane
1,2,3-Trichloropropane
1, 2, 4-Tri methyl benzene
1,3, 5-Tri methyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
Mean Accuracy
(% of True
Value, RSD
No.b 2 UQ/l Cone.) (%)
20
21



25
48
26
49
50
27
51
52
53
28
29
30
54
55
56
31
32
33
34
35
57
58
36
59
60
61
102
92



96
96
91
103
95
e
93
102
95
99
101
97
105
90
92
94'
107
99
81
97
93
88
104
97
f
98
2.2
3.7



1.7
9.1
5.3
3.2
3.6

7.6
4.9
4.4
2.7
4.6
4.5
2.8
5.7
5.2
3.9
3.4
2.9
4.6
3.9
3.1
2.4
3.5
1.8

2.3
Mean Accuracy
(% of True
Value, RSD
0.2 uq/L Cone.) (%)
103
93



99
100
88
101
95
e
78
97
104
95
84
92
126
78
83
94
109
106
48
91
106
97
115
98
f
103
2.9
3.2



2.1
4.0
2.4
2.1
3.1

8.3
2.1
3.1
3.8
3.6
3.3
1.7
2.9
5.9
2.5
2.8
2.5
13.
2.8
2.2
"3.2
14.
1.7

1.4
aData obtained using column 2 with the open split interface and an ion
 trap mass spectrometer (Sect. 11.3.2) with all method analytes in the same
 reagent water solution.
Designation  in Figures 1  and 2.
cNot measured; authentic standards were not available.
dNot found at 0.2 /jg/L.
eNot measured; methylene chloride was in the laboratory reagent blank.
fm-xylene coelutes with and cannot be distinguished from its isomer p-xylene,
No 61.
                                   524.2-41

-------
      TABLE 7.  ACCURACY AND PRECISION DATA FROM SEVEN DETERMINATIONS
                 OF METHOD  ANALYTES IN REAGENT WATER USING  WIDE  BORE
                 CAPILLARY  COLUMN NUMBER  4a
Compound
Acetone
Acrylonitrile
Ally! chloride
2-Butanone
Carbon disulfide
Chloroacetonitrile
1-Chlorobutane
t-Dichloro-2-butene
1, 1-Dichloropropanone
c-l,3-Dichloropropene
t-l,3-Dichloropropene
Di ethyl ether
Ethyl methacrylate
Hexachloroethane
2-Hexanone
Methacrylonitrile
Methyl acryl ate
Methyl iodide
Methyl methacryl ate
4-Methyl -2-pentanone
Methyl -tert-butyl ether
Nitrobenzene
2-Nitropropane
Pentachl oroethane
Propionitrile
Tetrahydrofuran
True
Cone.
(fig/L)
1.0
1.0
1.0
2.0
0.20
1.0
1.0
1.0
5.0
0.20
0.10
1.0
0.20
0.20
1.0
1.0
1.0
0.20
1.0
0.40
0.40
2.0
1.0
0.20
1.0
5.0
Mean
Cone.
Detected
(ug/L)
1.6
0.81
0.90
2.7
. 0.19
0.83
:0.87
1.3
4.2
0.20
0.11
0.92
;0.23
0.18
1.1
0.92
1.2
0.19
1.0
0.56
b.52
2.1
0.83
0.23
0.87
3.9
Rel.
Std.
Dev.
5.7
8.7
4.7
5.6
15
4.7
6.6
8.7
7.7
3.1
14
9.5
3.9
10
12
4.2
12
3.1
13
9.7
5.6
18
6.2
20
5.3
13
Method
Detect.
Limit
(M9/L)
0.28
0.22
0.13
0.48
0.093
0.12
0.18
0.36
1.0
0.020
0.048
0.28
0.028
0.057
0.39
0.12
0.45
0.019
0.43
0.17
0.090
1.2
0.16
0.14
0.14
1.6
Data obtained using column 4 with the open split interface and an ion trap
mass spectrometer.
                                 524.2-42

-------







a
o
u_ ^
°0

•G
ofc
1-4 "*
fe =
z ^
c **-
LU £-
££ uj
•?V fu
J""* |B^
•— Q£
0^
uf ^
z ";
^••^ «^
to *
^^ UJ
|i
a z
I1-1
ACCURACY ,
ANALYTES
co
UJ
_J
CO
«c
















CO
3

i— "aT
3
4- r-
OfTf
« w
QJ; S_x
UJ
1—
^5 > "ii*^*
rt l "vp
U*' ^*
a. a —
C —1
O) a>
0)
^3 ^' '^
S- a)
1— 3
f.
<4- r«
O 1^>
o; ^
i t i *^*
i
<
.^ •
2c OJ ^
C —1
us ~~~~
OJ CD
21 ^
O)

1— 
UJ S^
1— > -—
"ZL O ^
UJ C3 >— •
CJ3
UJ
o;
c: — J
O) CD










T3
C
3
O
a.
e
o
o





^5 ^^ ^^ &5 ^^ ^^ &^ &5 s^ ^? s^ s^
f^t ' if) 1/5 iff t°^ LO LO LO LO f^ t*^ LO
•— «ootoocri.i-HOOooi— ••— JcVl«^-KfSDLO
cocooNjr-^evj^tMCsjiococsjcsj

,— H C\4 ^D O^ CO CO O^ CSJ CNJ ^J CO *~^
cvJc\je\j^-tt-HOJt— ic\jc\jc\je\jc\j




Sootni-oooinoooo
CT* ^3 ^^D OO O^ ^D C3^ O** C5 O^ ^D ^D
r— ( r— I •— 1 f-H r-H i— t
1"**. i-H ^" i-H ^^ i-H VO I*"* h*» t— *
c\j«i-LOi-Hio»a-vo«a-LO 10 co 10
r-H f-H
f-H  C
3 0
eu JD c cu
1 — 1 t« 4->
aj -r- CM o. 
•as- i o . r— c
a> -i— +-> >ni
r— -i— r— CO+->
•f— S— 3 O +•* f-" S— <^" - 'JC O)O
1) = (J C nSS-CDUr— Si —
co ' n> c o o i •>- >> J=
Or— r— 4-» O S- r— CsJ Q J= r- O
4->>)>l3JDOJ= - I 4J >> rO
O) S- r— CQ S- i— O r-H f— I ,
c s-
ITJ O
x ca
cu -c:
ni 4J
1 CU



&5
LO
O
&?
f— 1
exj


S^
O
f-H


Q
£H
CO

CSJ
esj




o
r—i
f-H
^.
CO
.CSJ









O)
>
4->
CU
524.2-43

-------





















,— ,
-a
0)
3

,|_>
C
o
o
CO
LU
m
g


























CU
3
t— "oT
3
O OS

LU
cu s?
0. 0 — -
?
c: 'Hi
0$ 	 .
cu cn
S =1
**~^

cu
s- "oT
t— 3
M- 'oj
O >
LU 	
 ^-^
3 cu 55
eC 0 •— -
oc


^"^

cu cn
S .5


cu
H— cu
l*-r—
O OS
LU 55
)— •-— •
<
^ .
I— > ^-«
*£ cu ^5
cu
LU
^^
(—• 1
^_ '
cu cn
s:^.











-o
c
3
O
Q.
O
CH)




S5 5S S5 SS
in LO o o
en i— i i— i i— i


O 1 — CM U3
CO CM l> CO


en co CM CM
^H CM CM CM




5*5 &^ ^^ ^5
LO LO LO C)
en i — i o i— <

CO CO LO LO
co co in CM




en co .— i CM
t— 1 CM CM CM




v^
o o en a?
i— i , — i


^P ^g
^f ?~. ?~. LO
"* CO CO CO




CD CD cn en
CM CM .— 1 ,— (




S-
cu
CU -£=
C 4->
CU O CU
4-> C i —
OS 05 >,
^— 4-* 4-5
>> C 3
CU S- CU ^3
•O (J Q. 1
•i- 05 | 4->
-0 .£= CM S-
O 4-1 1 CU
-i- CU I— 4->
•— ^ £ ri

^) ^) 4**^ ^*)
•C Jd CU <^
cu cu i cu


s?
LO
CD


^
CM


1 — 1
CM




v^
CD
i — i

CO
•*




CM
CM





g
r— |



5f
LO




CD
CM











CU
enzen
t*^
Q
i.
£


^
o
1—1


CM
CO


CM
CM




xo
LO

»— (
If)




CO
CM





g
^H


0
i — 1
10




CD
CM











CU
C
03
Q.
O
s-
Q.
O
s_
4_9
•r~
1


&?
O
' — 1


; s
>-<


CM
CM




i, vp
LO
CD

SD
CM




i—H
i c\j

'



en




CM
LO




|2






i.


cu
|o3
,4->
cu
o
o
^
OS
4-J
C
CU


S5
LO
i— (


*
CM


CO
CM




v^
LO
i — (

en
CO




CO
CM





0
r— I



LO
*




CD
CM











CU
s-
4->
C
O
*,_«
Q.
i-


•^5
LO
O


en
CM


i — 1
CM




v,g
CD
CM

CM
CO




«s-
CM





0
1 — 1



cb
CM




CM










e—
S-
M-
O
•o
^
OS

cu
1



r—
1v
•£
'*
CU
cu
o
u
cu
CL
00
01
OS

Q.
c
"'—
03
•a
c:
05

CU
03 , •
f 1 ^_^
5- r-H
CU ' — ^
4->
•r- O
• (_
•i~~ ^3

a. o
co 01
£= S-
aj cu
O OS
cu
4-» £T
CU
-c cn
4-> OS
•r— CU
3 S-
«* CU
c: o$
3
r— CU
O -C
(J +J

cn c
C -r—

CO CO
3 CU

•0 >i
CU r—
C fO
03 03
-Q CO
O
CU
OS I—-

OS OS
Q H-
                  <0
524.2-44

-------
   FOAM
KW.
0. D. BUT
                  ton o. &.
                 VCUTHM.
                     0. 0.
                               YAlVf
                 r*-tic».» GAUGE
i. o. o. «jssa saiuu

     0. 0.   1/1( W. 0.0.
                                   SJEVf PURGE
                                   OASRLTBI
  IGUfl GUSS RB7
  HHXUII POROSiTt
                                     aot
                                     ccmwx
           FIGURE 1.  PURGING DEVICE
                    524.2-45

-------
     PACKING PftOCffiUJE

    &?«*
ACTIVATE), ,J
OtA*COAL.7.7
TUSINQ 2SCM
1105 Ui 10.
0*128 IN. 0.0.
                      souo
                (SINCU UTCQ
                       IOH
               mur
    FIGURE 2.  TRAP PACKINGS AND CONSTRUCTION TO INCLUDE
              OESORB CAPABILITY
                  524.2-46

-------
 't
1
3
C

-------
 i
 K
IT
i?
  Sf
 I.
                      524.2-48

-------
METHOD 525.2
DETERMINATION OF ORGANIC COMPOUNDS IN DRINKING WATER
BY LIQUID-SOLID EXTRACTION AND CAPILLARY COLUMN GAS
CHROMATOGRAPHY/MASS SPECTROMETRY
                            Revision 1.1
     J.W.  Eichelberger, T.D.  Behymer,  W.L.  Budde - Method 525,
     Revision 1.0, 2.0, 2.1 (1988)
     J.W.  Eichelberger,  T.D.  Behymer,  and W.L.  Budde - Method 525.1
     Revision 2.2 (July 1991)
     J.W.  Eichelberger,  J.W.  Munch,  and J.A.  Shoemaker
     Method 525.2  Revision 1.0 (February,  1994)

     J.W.  Munch - Method 525.2,  Revision 2.0  (1995)
               NATIONAL EXPOSURE RESEARCH LABORATORY
                 OFFICE  OF  RESEARCH  AND  DEVELOPMENT
               U.S.  ENVIRONMENTAL PROTECTION AGENCY
                      CINCINNATI, OHIO  45268
                              525.2-1

-------
                                 METHOD 525.2

             DETERMINATION OF ORGANIC COMPOUNDS IN DRINKING WATER
               BY LIQUID-SOLID EXTRACTION AND CAPILLARY COLUMN
                     GAS CHROMATOGRAPHY/MASS SPECTROMETRY
1.   SCOPE AND APPLICATION

     1.1  This is a general purpose method that provides procedures for
          determination of organic compounds in finished drinking water,
          source water, or drinking water in any treatment stage.  The method
          is applicable to a wide range of organic compounds that are
          efficiently partitioned from the water sample onto a C18 organic
          phase chemically bonded to a solid matrix in a disk or cartridge,
          and sufficiently volatile and thermally stable for gas chromatog-
          raphy.  Single-laboratory accuracy and precision data have been
          determined with two instrument systems using both disks and car-
          tridges for most of the following compounds:
           Analvte

          Acenaphthylene
          Alachlor
          Aldrin
          Ametryn
          Anthracene
          Atraton
          Atrazine
          Benz[a]anthracene
          Benzo[b]fluoranthene
          Benzo[k]fluoranthene
          Benzo[a]pyrene
          Benzo[g,h,i]perylene
          Bromacil
          Butachlor
          Butyl ate
          Butyl benzylphthalate
          Carboxin2
          Chlordane components
            Alpha-chlorda.ne
            Gamma-chlordane
            Trans nonachlor
          Chlorneb
          Chlorobenzilate
          Chlorpropham
          Chlorothalonil
          Chlorpyrifos
          2-Chlorobiphenyl
MW
 152
 269
 362
 227
 178
 211
 215
 228
 252
 252
 252
 276
 260
 $11
 217
 312
 235

 406
 406
 440
 206
 324
 213
 264
 349
 188
Chemical  Abstracts Service
     Registry Number	
           208
         15972
           309
           834
           120
          1610
          1912
            56
           205
           207
            50
           191-
           314-
         23184-
          2008-
            85-
          5234-

          5103-
          5103-
         39765-
          2675-
           510-
           101-
          1897-
          2921-
          2051-
•96-8
-60-8 '
-00-2
-12-8
-12-7
-17-9
-24-9
-55-3
-82-3
-08-9
-32-8
-24-2
-40-9
-66-9
-41-5
-68-7
-68-4

 71-9
 74-2
 80-5
 77-6
 15-6
 21-3
 45-6
 88-2
 60-7
                                    525.2-2

-------
 Chrysene                         228
 Cyanazine                        240
 Cycloate                         215
 Dacthal(DCPA)                    330
 ODD, 4,4'-.                      318
 DDE, 4,4'-                       316
 DDT, 4,4'-                       352
 Diazinon                         304
 Dibenz[a,h]anthracene            278
 Di-n-butylphthalate              278
 2,3-Dichlorobiphenyl             222
 Dichlorvos                       220
 Dieldrin                         378
 Diethylphthalate                 222
 Di(2-ethylhexyl)adipate          370
 Di(2-ethylhexyl,)phthalate        390
 DimethylIphthalate                194
 2,4-Dinitrotoluene               182
 2,6-Dinitrotoluene               182
 Diphenamid                       239
 Disulfoton2                      274
 Disulfoton sulfoxide2         •   290
 Disulfoton sulfone              306
 Endosulfan I                    404
 Endosulfan II                   404
 Endosulfan sulfate              420
 Endrin                           373
 Endrin  aldehyde                 378
 EPIC                            189
 Ethoprop                        242
 Etridiazole                     246
 Fenamiphos2                      303
 Fenarimol       .                 330
 Fluorene                        166
 Fl undone                        328
 Heptachlor                       370
 Heptachlor  epoxide              386
 2,2',3,3',4,4',6-Heptachloro-
   biphenyl                       392
 Hexachlorobenzene                282
 2,2',4,4',5,6'-Hexachloro-
   biphenyl                       353
 Hexachlorocyclohexane,  alpha     288
 Hexachlorocyclohexane,  beta      288
 Hexachlorocyclohexane,  delta     288
 Hexachlorocyclopentadiene        270
 Hexazinone                       252
 Indeno[l,2,3,c,d]pyrene          276
 Isophorone                       133
 Lindane                          288
Merphos                          298
Methoxychlor                     344
   218
 21725
  1134
  1861
    72
    72
    50
   333
    53
    84-
 16605-
    62-
    60-
    84-
   103-
   117-
   131-
   121-
   606-
   957-
   298-
  2497-
  2497-
   959-
 33213-
  1031-
    72-
  7421-
   759-
 13194-
  2593-
 22224-
 6016,8-
   86
 59756
   76
  1024
  01-9
-46-2
-23-2
-32-1
-54-8
-55-9
-29-3
-41-5
-70-3
-74-2
-91-7
-73-7
-57-1
-66-2
-23-1
-81-7
-11-3
-14-2
-20-2
-51-7
-04-4
-07-6
-06-5
-98-8
-65-9
-07-8
-20-8
-93-4
•94-4
•48-4
 15-9
 92-6
 88-9
 73-7
 60-4
 44-8
 57-3
52663-71-5
  118-74-1
60145-
  319-
  319-
  319-
   77-
51235-
  193-
   78-
   58-
  150-
   72-
-22-4
•84-6
•85-7
•86-8
•47-4
04-2
39-5
59-1
89-9
50-5
43-5
                          525.2-3

-------
Methyl paraoxon
Metolachlor
Metribuzin
Mevinphos
MGK 264
Molinate
Napropamide
Norflurazon
2,2',3,3',4,5',6,6'-Octa-
chlorobiphenyl
Pebulate
2,2',3',4,6-Pentachloro-
biphenyl
Pentachlorophenol
Phenanthrene
Permethrin, cis-
Permethrin, trans
Prometon
Prometryn
Pronamide
Propachlor
Propazine
Pyrene
Simazine
Simetryn
Stirofos
Tebuthiuron
Terbacil
Terbufos2
Terbutryn
2,2' ,4,4'-Tetrach1orobiphenyl
Toxaphene
Triademefon
2,4,5-Trichlorobiphenyl
Tricyclazole
Trifluralin
Vernolate
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
247
283
214
224
275
187
271
303

426
203

324
264
178
390
390
225
241
255
211
229
202
201
213
364
228
216
288
241
290

293
256
189
335
203





;

950-35-6
51218-45-2
21087-64-9
7786-34-7
113-48-4
2212-67-1
15299-99-7
27314-13-2

40186-71-8
1114-71-2

60233-25-2
87-86-5
85-01-8
54774-45-7
51877-74-8
1610-18-0
7287-19-6
23950-58-5
1918-16-7
139-40-2
129-00-0
122-34-9
1014-70-6
22248-79-9
34014-18-1
5902-51-2
13071-79-9
886-50-0
2437-79-8
8001-35-2
43121-43-3
15862-07-4
41814-78-2
1582-09-8
1929-77-7
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
^onoisotopic molecular  weight  calculated from the atomic masses of
the isotopes with the smallest  masses.
                         525.2-4

-------
            Only qualitative identification  of these  analytes  is  possible
           because of their instability in aqueous matrices.   Merphos,  car-
           boxin, disulfoton, and disulfoton sulfoxide showed  instability
           within 1 h of fortification.  Diazinon, fenamiphos, and terbufos
           showed significant losses within  7 days under the  sample storage
           conditions specified in this method;

           Attempting to determine all  of the above analytes  in all  samples is
           not practical and not necessary in most cases.   If  all  the analytes
           must be determined,  multiple calibration mixtures will  be required.

      1.2   Method detection limit (MDL) is defined as the  statistically calcu-
           lated minimum amount that can be  measured  with  99%  confidence that
           the reported  value is greater than zero (1).  The MDL  is  compound
           dependent  and is particularly dependent on extraction  efficiency and
           sample matrix.   MDLs for all method analytes  are listed in Tables 3
           through 6.  The  concentration calibration  range demonstrated in  this
           method is  0.1 //g/L to 10 ng/L for most  analytes.

2.    SUMMARY  OF METHOD

      Organic  compound analytes,  internal  standards,  and surrogates  are
      extracted  from  a water sample by  passing 1 L of sample water through  a
      cartridge  or disk  containing  a solid matrix  with a chemically  bonded  C18
      organic  phase (liquid-solid  extraction,  LSE).   The organic compounds  are
      eluted from the LSE cartridge or  disk  with small quantities  of ethyl
      acetate  followed by methylene chloride,  and  this extract  is  concentrated
      further  by evaporation of some of the  solvent.  The  sample components are
      separated,  identified,  and measured  by  injecting an  aliquot  of the
      concentrated  extract  into  a  high  resolution  fused silica  capillary column
      of a gas chromatography/mass  spectrometry (GC/MS) system.  Compounds
      eluting  from  the GC column are  identified by  comparing their measured
      mass spectra  and retention times  to  reference spectra and retention times
      in a data  base.  Reference spectra and  retention times for analytes are
      obtained by  the measurement of calibration standards under the  same
      conditions  used for samples.   The concentration of each  identified
      component  is  measured  by relating the MS response of the  quantitation ion
      produced by that compound to  the  MS  response of the quantitation  ion
      produced by a compound that is  used  as  an internal  standard.  Surrogate
      analytes, whose concentrations  are known in every sample, are measured
     with the same internal standard calibration procedure.

3.   DEFINITIONS

     3.1  INTERNAL STANDARD (IS) — A  pure analyte(s) added to a sample,
          extract, or standard solution in known amount(s) and used to measure
          the relative responses of other method analytes  and  surrogates that
          are components of the same solution.  The internal  standard must be
          an analyte  that is not a sample component.

     3.2  SURROGATE ANALYTE (SA) - A pure analyte(s), which  is extremely
          unlikely to be found in any sample, and  which  is added to a sample

                                    525.2-5

-------
     aliquot in known amount(s) before extraction or other processing,
     and is measured with the same procedures used to measure other
     sample components.  The purpose of the SA is to monitor method
     performance with each sample.

3.3  LABORATORY DUPLICATES (LD1 and LD2) — Two aliquots of the same
     sample taken in the laboratory and analyzed separately with iden-
     tical procedures.  Analyses of LD1 and LD2 indicate precision
     associated with laboratory procedures, but not with sample collec-
     tion, preservation, or storage procedures.

3.4  FIELD DUPLICATES (FD1 and FD2) — TWO separate samples collected at
     the same time and place under identical circumstances, and treated
     exactly the same throughout field and laboratory procedures.
     Analyses of FD1 and FD2 give a measure of the precision associated
     with sample collection, preservation, and storage, as well as with
     laboratory procedures.
                       i                 I,
3.5  LABORATORY REAGENT BLANK  (LRB) — An aliquot of reagent water or
     other blank matrix that is treated exactly as a sample including
     exposure to all glassware, equipment, solvents, reagents, internal
     standards, and surrogates that are used with other samples.  The LRB
     is used to determine if method analytes or other interferences are
     present in the laboratory environment, the reagents, or the
     apparatus.

3.6  FIELD REAGENT BLANK (FRB) — An aliquot of reagent water or other
     blank matrix that is placed  in a sample container  in the laboratory
     and treated as a sample in all respects,  including shipment to the
     sampling site, exposure to sampling  site  conditions, storage,
     preservation, and all analytical procedures.  The  purpose of the FRB
     is to determine  if method analytes or other  interferences are
     present in the freld environment.

3.7  INSTRUMENT PERFORMANCE CHECK SOLUTION  (IPC)  — A solution of one or
     more method analytes, surrogates,  internal standards, or other test
     substances used  to evaluate  the performance  of the instrument system
     with respect to  a defined set of method criteria.

3.8  LABORATORY FORTIFIED BLANK (LFB) ---  An aliquot of  reagent water or
     other blank matrix to which  known  quantities of the method analytes
     are  added  in the laboratory.  The  LFB  is  analyzed  exactly like a
     sample, and its  purpose is to determine whether the methodology is
     in control, and  whether the  laboratory is capable  of making accurate
     and  precise measurements.

3.9  LABORATORY FORTIFIED SAMPLE  MATRIX  (LFM>  —  An aliquot of an
     environmental sample to which known  quantities of  the method
     analytes are added  in the laboratory.  The LFM is  analyzed exactly
     like a sample, and  its purpose  is  to determine whether the sample
     matrix contributes  bias to the  analytical results.  The  background
     concentrations of  the analytes  in  the  sample matrix must  be

                               525.2-6

-------
           determined in a separate aliquot and the measured values in the LFM
           corrected for background concentrations.

      3.10 STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing
           one or more method analytes prepared in the laboratory using assayed
           reference materials or purchased from a reputable commercial source.

      3.11 PRIMARY DILUTION STANDARD SOLUTION (PDS) — A solution of several
           analytes prepared in the laboratory from stock standard solutions
           and diluted as needed to prepare calibration solutions and other
           needed analyte solutions.

      3.12 CALIBRATION STANDARD (CAL)  -- A solution prepared from the primary
           dilution standard solution  or stock standard solutions and the
           internal standards and surrogate analytes.   The CAL  solutions  are
           used to calibrate the instrument response with respect to analyte
           concentration.

      3.13 QUALITY CONTROL  SAMPLE (QCS)  - A solution  of method  analytes  of
           known  concentrations  which  is used  to  fortify an aliquot  of LRB  or
           sample matrix.   The  QCS  is  obtained from a  source external  to  the
           laboratory  and different  from the source of calibration  standards.
           It  is  used  to  check  laboratory performance  with externally  prepared
           test materials.

4.    INTERFERENCES

      4.1   During  analysis, major contaminant  sources  are  reagents and  liquid-
           solid  extraction devices.  Analyses of field  and  laboratory  reagent
           blanks  provide information about  the presence of  contaminants.

      4.2   Interfering contamination may  occur when  a  sample containing low
           concentrations of  compounds is  analyzed  immediately after a  sample
           containing relatively high concentrations of compounds.  Syringes
           and  splitless injection port liners must  be cleaned carefully or
           replaced as needed.  After analysis of a  sample containing high
           concentrations of compounds, a laboratory reagent blank should be
           analyzed to ensure that accurate values are obtained for the next
           sample.

5.   SAFETY

     5.1  The toxicity or carcinogenicity of chemicals used in  this method has
          not been precisely defined;  each chemical should be treated as  a
          potential health  hazard,  and exposure to these chemicals should be
          mlnoo,!«ed-   Fach  1ab°ratory  is responsible for maintaining awareness
          of OSHA regulations regarding  safe handling of chemicals used in
         .this method.   Additional  references  to  laboratory safety are cited


     5.2  Some method  analytes  have  been tentatively classified  as  known  or
          suspected human or mammalian carcinogens.   Pure standard  materials
          and  stock standard  solutions of these compounds  should  be  handled
          with suitable  protection  to  skin,  eye's,  etc.

                                   525.2-7

-------
6.   EQUIPMENT AND SUPPLIES (All specifications are suggested.  Catalog
     numbers are included for illustration only.)
     6.1
     6.2
     All  glassware must be meticulously cleaned.   This  may be
     accomplished by washing with detergent and water,  rinsing with
     water,  distilled water, or solvents,  air-drying,  and heating (where
     appropriate) in a muffle furnace.   Volumetric glassware should never
     be heated to the temperatures obtained in a muffle furnace.

     Sample containers.  1-L or 1-qt amber glass bottles fitted with
     Teflon-lined screw caps.  Amber bottles are highly recommended since
     some of the method analytes are very sensitive to light and are
     oxidized or decomposed upon exposure.

6.3  Volumetric flasks, various sizes.
                                                                 s
6.4  Laboratory or aspirator vacuum system.  Sufficient capacity to
     maintain a minimum vacuum of approximately 13 cm (5 in.) of mercury
     for cartridges.  A greater vacuum (66 cm  [26 in.] of mercury)  may be
     used with disks.

6.5  Micro syringes, various sizes.

6.6  Vials.  Various sizes  of amber vials with Teflon-lined  screw caps.

6.7  Drying column.  The drying tube should contain about 5  to 7 grams of
     anhydrous sodium  sulfate to prohibit residual water from
     contaminating the extract.  Any small tube may be used,  such as a
     syringe barrel, a glass dropper,  etc. as  long as no sodium  sulfate
     passes through  the column  into the extract.

6.8  Analytical  balance.   Capable of weighing  0.0001 g accurately.

6  9  Fused silica  capillary gas  chromatography column.   Any  capillary
     column that provides  adequate  resolution,' capacity,  accuracy,  and
     precision  (Sect.  10)  can  be  used.  Medium polar,  low bleed  columns
     are  recommended for  use with this method  to  provide adequate
     chromatography  and minimize  column bleed.  A 30 m X 0.25 mm id fused
     silica capillary  column coated with  a  0.25 /zm bonded film of
     polyphenylmethylsilicone  (J&W  DB-5.MS) was used to  develop  this
     method.  Any column  which  provides  analyte separations  equivalent  to
     or  better  than  this  column may be used.

6.10 Gas  chromatograph/mass spectrometer/data system  (GC/MS/DS).

     6 10.1 The  GC must  be capable  of  temperature programming and  be
             equipped for  splitless/split  injection.   On-column capillary
             injection  is  acceptable if all  the quality control
             specifications in  Sect. 9  and Sect.  10 are met.   The
             injection  tube liner should be quartz and  about 3 mm in
             diameter.   The injection  system must  not  allow the analytes
                                     525.2-8

-------
                  to contact hot stainless steel or other metal surfaces that
                  promote decomposition.

           6.10.2 The GC/MS interface should allow the capillary column or
                  transfer line exit to be placed w.ithin a few mm of the ion
                  source.  Other interfaces, for example the open split inter-
                  face, are acceptable as long as the system has adequate
                  sensitivity (see Sect. 10 for calibration requirements).

           6.10.3 The mass spectrometer must be capable of electron
                  ionization at a nominal electron energy of 70 eV to produce
                  positive ions.  The spectrometer must be capable of scanning
                  at a minimum from 45 to 450 amu with a complete scan cycle
                  time (including scan overhead) of 1.0 sec or less.   (Scan
                  cycle time = total  MS data acquisition time in sec  divided by
                  number of scans in  the chromatogram).   The spectrometer must
                  produce a mass spectrum that meets  all  criteria in  Table  1
                  when an injection of approximately  5 ng of DFTPP is
                  introduced into the GC.   An average spectrum across the DFTPP
                  GC peak may be used to test instrument performance.  The  scan
                  time should be set  so that all  analytes have a minimum of 5
                  scans across  the  chromatographic  peak.

           6.10.4 An interfaced  data  system is  required  to  acquire, store,
                  reduce,  and  output  mass  spectral  data.   The computer software
                  must have  the  capability  of processing  stored  GC/MS data  by
                  recognizing  a  GC  peak within  any  given  retention  time  window,
                  comparing  the  mass  spectrum from  the GC peak with spectral
                  data in  a  user-created  data base, and  generating  a  list of
                  tentatively  identified  compounds  with  their retention  times
                  and  scan numbers.   The  software must also  allow  integration
                  of the  ion  abundance  of any specific ion between  specified
                  time  or  scan number limits, calculation of  response factors
                  as defined  in  Sect.  10.2.6  (or construction  of a  linear
                  regression calibration curve), calculation  of response factor
                  statistics  (mean  and  standard deviation), and calculation of
                  concentrations of analytes  using  either the  calibration curve
                  or the equation in  Sect.  12.

     6.11 Standard  Filter Apparatus, ALL GLASS OR TEFLON LINED.  These should
          be used to carry out disk  extractions when  no automatic system or
          manifold  is utilized.

     6.12 A manifold system or an automatic or robotic commercially  available
          sample preparation system  designed for either cartridges or disks
          may be utilized in this method if all quality control  requirements
          discussed in Sect. 9 are met.

7.   REAGENTS AND STANDARDS

     7.1  Helium carrier gas, as contaminant free as  possible.


                                   525,2-9

-------
7.2  Liquid-solid extraction (LSE) cartridges.  Cartridges are inert
     non-leaching plastic, for example polypropylene, or glass, and must
     not contain plasticizers, such as phthalate esters or adipates, that
     leach into the ethyl acetate and methylene chloride eluant.  The
     cartridges are packed with about 1 gram of silica, or other inert
     inorganic support, whose surface is modified by chemically bonded
     octadecyl (C18) groups.   The  packing must  have  a narrow  size
     distribution and must not leach organic compounds into the eluting
     solvent.  One liter of water should pass through the cartridge in
     about 2 h with the assistance of a slight vacuum of about 13 cm (5
     in.) of mercury.   Sect. 9 provides criteria for acceptable LSE
     cartridges which are available from several commercial suppliers.

     The extraction disks contain octadecyl bonded silica uniformly
     enmeshed in an inert matrix.  The disks used to generate the data in
     this method were 47 mm in diameter and 0.5 mm in thickness.  Other
     disk sizes are acceptable and larger disks may be used for special
     problems or when sample compositing "is carried out.  As with
     cartridges, the disks should not contain any organic compounds,
     either from the matrix or the bonded silica, which will leach into  .
     the ethyl acetate and methylene chloride eluant.  One liter of
     reagent water should pass through thja disks in 5-20 min using a
     vacuum of about 66 cm (26 in.) of mercury.  Sect. 9 provides
     criteria for acceptable LSE disks wh[ich are available commercially.

7.3  Solvents

     7.3.1  Methylene chloride, ethyl acetate, acetone, toluene and
            methanol.   High purity pesticjide quality or equivalent.

     7.3.2  Reagent water.  Water in which an interference is not
            observed at the method detection5 limit of the compound of
            interest.   Prepare reagent water by passing tap water through
            a filter bed containing about 0.5 kg of activated carbon or
            by using a water purification system.  Store in clean,
            narrow-mouth bottles with TefTon-lined septa and screw caps.

7.4  Hydrochloric acid. 6N.

7.5  Sodium sulfate, anhydrous.   (Soxhlet extracted with methylene
     chloride for a minimum of 4 h or heated to 400°C for 2 h in a muffle
     furnace.)            .

7.6  Stock standard solutions.  Individual solutions of surrogates,
     internal standards, and analytes, or mixtures of analytes, may-be
     purchased from commercial suppliers or prepared from pure materials.
     To prepare, add 10 mg (weighed on an analytical balance to 0.1 mg)
     of the pure material to 1.9 mL of methanol, ethyl acetate, or
     acetone in a 2-mL volumetric flask, dilute to the mark, and transfer
     the solution to an amber glass vial.  If the analytical standard  is
     availab-le only in quantities smaller than  10 mg, reduce the volume
     of solvent accordingly.  Some polycyclic aromatic hydrocarbons are

                              525.2-10   :

-------
                                                                                       I
      not  soluble  in methanol,  ethyl  acetate,  or  acetone,  and  their  stock
      standard  solutions  are  prepared in  toluene.  Methylene chloride
      should be  avoided as  a  solvent  for  standards because  its  high  vapor
      pressure  leads to rapid evaporation  and  concentration changes
      Methanol,  ethyl acetate,  and acetone  are not as volatile  as
      methylene  chloride, but their solutions must also be handled with
      care to avoid evaporation.  If  compound purity is confirmed by the
      supplier at  >96%, the weighed amount  can be used without  correction
      to calculate the concentration  of the solution (5 iiq/u,n   store the
      amber vials  at 4° C or less.

 7.7  Primary dilution standard solution.  The stock standard solutions
      are used to prepare a primary dilution standard solution  that
      contains multiple analytes.  Mixtures of these analytes to be used
      as primary dilution standards may be purchased from commercial
      suppliers.  Do not put every method analyte in a single primary
      dilution standard because chromatographic separation will be
      extremely difficult, if not impossible.   Two or three primary
      dilution standards would be more appropriate.   The recommended
      solvent for these standards is acetone or ethyl  acetate.   Aliquots
      of each of the stock standard solutions  are combined to produce the
      primary dilution  in  which the concentration of the analytes is  at
      least equal to the concentration of the  most concentrated
      calibration solution,  that is,  10 ng//il_.   Store  the  primary dilution
      standard  solution  in an  amber vial  at 4°  C  or  less,  and  check
      frequently for signs of  degradation  or evaporation,  especially  just
      before  preparing  calibration solutions.

 7.8   Fortification solution of internal  standards and  surrogates
      Prepare  an  internal  standard solution of  acenaphthene-D1fl
      phenanthrene-D10,  and chrysene-D12, in  methanol, ethyl  acetate,  or
      acetone  at  a  concentration of 500 ng/ml of  each.   This solution is
      used  in the preparation  of the calibration  solutions.  Dilute a
      portion of  this solution  by  10 to a  concentration  of  50 /ig/mL and
      use this solution  to fortify the actual water samples  (see Sect
     .11.1.3 and  Sect. 11.2.3).   Similarly,  prepare  both surrogate
      compound solutions (500/yg/mL for calibration, 50//g/mL for
      fortification).  Surrogate compounds  used in developing this method
      are l,3-dimethyl-2-nitrobenzene,  perylene-D12,  and
      triphenylphosphate.  Other surrogates, for example pyrene-D1n may be
      used  in this  solution as needed  (a 100-/iL aliquot of this  50 /ig/mL
      solution added to 1 L of water gives a concentration of 5  /zg/L  of
      each  internal standard or  surrogate).  Store these solutions in an
      amber vial   at 4° C or less.  These two solutions may be combined or
     made as a single solution.

7.9  GC/MS performance check solution.  Prepare a solution in methylene
     chloride of the following compounds at 5  ng//zL of each:  DFTPP and
     endrin,  and 4,4'-DDT.  Store this solution in an amber vial at 4° C
     or less.   DFTPP is less stable in acetone or ethyl  acetate than  it
     is in methylene chloride.                             '
                              525.2-1!

-------
    7  10  Calibration  solutions  (CAL1  through  CAL6).   Prepare  a  series of six
          concentration  calibration  solutions  in  ethyl  acetate which  contain
          analytes  of  interest (except pentachlorophenol,  toxaphene,  and the
          Aroclor compounds)  at  suggested  concentrations  of  10,  5,  2,  1, 0.5,
          and  0.1 ng//iL,  with a  constant concentration of 5  ng//iL  of  each
          internal  standard and  surrogate  in  each CAL solution.   It should  be
          noted that CAL1 through CAL6 are prepared by combining appropriate
          aliquots  of  a  primary  dilution standard solution (Sect.  7.7)  and  the
          fortification  solution (500  /ig/mL)  of internal  standards and
          surrogates (Sect. 7.8).  All calibration solutions should contain at
          least 80% ethyl acetate to avoid gas chromatographic problems.  IF
          ALL  METHOD ANALYTES ARE TO BE DETERMINED, TWO OR THREE SETS OF
          CALIBRATION  SOLUTIONS  WILL LIKELY BE REQUIRED.   Pentachlorophenol is
          included  in  this solution at a concentration four times the other
          analytes. Toxaphene CAL solutions  should be prepared as separate
          solutions at concentrations  of 250,  200, 100, 50,  25,  and 10 ng//iL.
          Aroclor CAL  solutions  should be  prepared individually at
          concentrations of 25,  10,  5, 2.5, 1., 0.5 and 0.2 ng/pl.  Store
          these solutions in amber vials in a dark cool place.  Check these
          solutions regularly for signs of degradation, for example,  the
          appearance of anthraquinone from the oxidation of anthracene.

     7.11 Reducing agent.  Sodium sulfite, anhydrous.  Sodium thiosulfate  is
          not recommended as it may produce a residue of elemental sulfur  that
          can interfere with some analytes.

     7 12 Fortification solution for  recovery standard.  Prepare a solution of
          terphenyl-Du  at a  concentration of  500 /ig/mL in methylene  chloride
          or ethyl  acetate.  These solutions are also commercially available.
          An aliquot of this solution should be  added to each extract to check
          on the recovery of the internal   standards .in the extraction process.

8.   SAMPLE COLLECTION. PRESERVATION. AND  STORAGE

     8.1  Sample collection.  When sampling from a water tap, open the tap and
          allow  the system to flush until   the water  temperature has  stabilized
          (usually about  2 min).  Adjust  the flow  to  about  500 mL/rhin and
          collect  samples  from  the flowing stream.   Keep  samples  sealed from
          collection time  until  analysis.  When  sampling  from an  open body of
          water, fill  the  sample container with  water  from  a  representative
          area.  Sampling  equipment,  including automatic  samplers, must be
          free of  plastic  tubing, gaskets, and other parts  that may  leach
          interfering analytes  into the water  sample.  Automatic  samplers  that
          composite samples  over time should  use refrigerated glass  sample
          containers  if  possible.

     8.2  Sample dechlorination and preservation.   All  samples  should be iced
          or  refrigerated at 4°C and  kept  in the dark  from the time of
          collection  until  extraction.  Residual chlorine should  be  reduced at
          the  sampling  site  by  addition of 40-50 mg of sodium sulfite  (this
          may  be added  as a  solid with  stirring  or shaking  until  dissolved) to
          each water  sample.   It is very  important that  the sample be

                                    525.2-12

-------
      dechlorinated prior to adding acid to lower the pH of the sample.
      Adding sodium sulfite and HC1 to the sample bottles prior to
      shipping to the sampling site is not permitted.  Hydrochloric acid
      should be used at the sampling site to retard the microbiological
      degradation of some analytes in water.  The sample pH is adjusted  to
      <2 with 6 N hydrochloric acid.  This is the same pH used in the
      extraction, and is required to support the recovery of acidic
      compounds like pentachlorophenol.                 •

      8.2.1  If cyanizine is to be determined,  a separate sample must be
             collected.   Cyanazine degrades in  the sample when it is
             stored under acidic conditions or  when sodium sulfite is
             present in  the stored sample.   Samples collected for
             cyanazine determination MUST NOT be dechlorinated or
             acidified when collected.   They should be iced or
             refrigerated as described  above and analyzed within 14  days.
             However,  these samples MUST be dechlorinated and acidified
             immediately prior to fortification with internal  standards
             and  surrogates,  and extraction using the same quantities of
             acid and  sodium sulfite described  above.

      8.2.2  Atraton and prometon are not efficiently extracted  from water
             at pH 2 due to what appears to be  their ionization  in
             solution  under acidic conditions.   In  order  to determine
             these analytes accurately,  a separate  sample must be
             collected and  dechlorinated with sodium sulfite,  but  no acid
             should  be added.   At neutral  pH, these  two compounds  are
             recovered from water with efficiencies  greater than 90%.  The
             data  in Tables  3,  4,  5,  6,  and 8 are  from  samples extracted
             at pH 2.

8.3   Holding time.  Results of the  time/storage  study of all method
      analytes showed  that  all  but  six compounds  are  stable  for  14 days  in
      water  samples  when  the samples  are  dechlorinated, preserved, and
      stored  as described in Sect.  8.2.   Therefore,  samples  must be
      extracted within 14 days.   If  the  following analytes  are to be
      determined,  the  samples cannot  be  held for  14 days  but must be
      extracted immediately after  collection and preservation: carboxin,
      diazinon, disulfoton, disulfoton sulfoxide, fenamiphos, and
      terbufos.  Sample  extracts may  be  stored  at 4° C for  up to 30 days
      after sample extraction.

8.4   Field blanks.

     8.4.1  Processing of a field reagent  blank (FRB) is recommended
            along with each sample set,  which  is composed of the samples
            collected from the same general sample site at approximately
            the same time.   At the laboratory,  fill a sample container
            with reagent water, seal, and ship  to the sampling site,along
            with the empty sample containers.  Return the FRB to the
            laboratory with the filled sample bottles.


                              525.2-13

-------
          8.4.2  When sodium sulfite and hydrochloric acid are added to
                 samples, use the same procedure to add the same amounts to
                 the FRB.                   ;

9.   QUALITY CONTROL

     9.1  Quality control (QC) requirements are the initial demonstration of
          laboratory capability followed by regular analyses of laboratory
          reagent blanks, laboratory fortified blanks, and laboratory
          fortified matrix samples.  A MDL should be determined for each
          analyte of interest.  The laboratory must maintain records to
          document the quality of the data generated.  Additional quality
          control practices are recommended.:

     9.2  Initial demonstration of low disk ()r cartridge system background.
          Before any samples are analyzed, or any time a new supply of
          cartridges or disks is received from a supplier, it must be demon-
          strated that a laboratory reagent blank (LRB) is reasonably free of
          contamination that would prevent t\\e determination of any analyte of
          concern.  In this same experiment,: it must be demonstrated that the
          particle size and packing of the LSE cartridges or the preparation
          of the disks are acceptable.  Consistent flow rate with all samples
          is an indication of acceptable particle size distribution, packing,
          and proper preparation.

          9.2.1  A source of potential contamination is the liquid-solid
                 extraction (LSE) cartridge or disk which could contain
                 phthalate esters, silicon compounds, and other contaminants
                 that could prevent the determination of method analytes (5).
                 Although disks are generally made of an inert matrix, they
                 may still contain phthalate material.  Generally, phthalate
                 esters can be leached from the cartridges into ethyl acetate
                 and methylene chloride and produce a variable background in
                 the water sample.  If the background contamination is
                 sufficient to prevent accurate and precise measurements, the
                 condition must be corrected before proceeding with the
                 initial demonstration.

          9.2.2  Other sources of background contamination are solvents,
                 reagents, and glassware.  Background contamination must be
                 reduced to an acceptable level before proceeding with the
                 next section.  In general, background from method analytes
                 should be below the method detection limits.

          9.2.3  One liter of water should pass through a cartridge in about 2
                 h with a partial vacuum of about 13 cm (5 in.) of mercury.
                 Using full aspirator or pump vacuum, approximately 5-20 min
                 will normally be required t;o pass one liter of drinking water
                 through a disk.  The extraction time should not vary
                 unreasonably among LSE cartridges or disks.
                                   525.2-14

-------
 9.3
9.4
      9.3.2
      9.3,3
  Initial demonstration of  laboratory accuracy and precision.  Analyze
  four to seven replicates  of a laboratory fortified blank containing
  each analyte of concern at a suggested concentration  in the range of
  2-5 /jg/L.  This concentration should be approximately in the middle
  of the calibration range, and will be dependent on the sensitivity
  of the instrumentation used.

  9.3.1  Prepare each replicate by adding sodium sulfite and HC1
        according to Sect. 8.2, then adding an appropriate aliquot of
        the primary dilution standard solution,  or  certified quality
        control sample, to reagent water.   Analyze each replicate
        according to the procedures described in Sect. 11.

        Calculate the measured concentration of each analyte in each
        replicate,  the mean concentration  of each analyte in alf
        replicates,  and mean accuracy (as  mean percentage of true
        value) for each analyte,  and the precision (as relative
        standard deviation, RSD)  of the measurements for each
        analyte.      .       .

        For each analyte and surrogate,  the  mean accuracy,  expressed
        as  a percentage of the  true value,  should be 70-130% and the
        RSD should  be <30%.  If these criteria are not met,  locate
        the source  of the  problem,  and  repeat with freshly oreoared
        LFBs.                   •

        Analyze  seven replicate  laboratory  fortified blanks  which
        have been  fortified with  all  analytes of interest  at
        approximately 0.5  /ig/L.   Calculate the MDL of  each  analyte
        using  the procedure described  in Sect. 13.1.2  (1).   It  is
        recommended  that these  analyses  be performed over  a  period of
        three  or  four days  to produce more realistic method  detection
        limits.

 9.3.5   Develop  and  maintain a system of control  charts to plot  the
        precision and  accuracy of analyte and  surrogate measurements
        as  a function  of time.  Charting of  surrogate  recoveries is
        an  especially  valuable activity since  these  are present  in
        every  sample  and the analytical results will form a
        significant  record  of data quality.

Monitor the integrated areas of the quantitation  ions of the
 internal standards  and surrogates in continuing .calibration checks
 (see Sect.  10.3).   In laboratory fortified  blanks or  samples, the
 integrated  areas of  internal standards and  surrogates will not be
constant because the volume of the extract  will  vary  (and is
difficult to keep constant).  But the ratios of the areas should be
reasonably  constant in laboratory fortified blanks and samples.  The
addition of 10 fj.1 of the recovery standard,  terphenyl-D14  (500
/ig/mL),  to  the extract is  recommended to be used to monitor the
recovery of the internal standards in laboratory fortified blanks
and samples.  Internal standard recovery should be in excess of 70%.
     9.3.4
                              525.2-15

-------
9.5
9.6
9.7
9.8
With each batch of samples processed as a group within a 12 h work
shift, analyze a laboratory reagent blank to determine the
background system contamination.  Any time a new batch of LSE
cartridges or disks is received, or new supplies of other reagents
are used, repeat the demonstration of low background described in
Sect. 9.2.                         !

With each batch of samples processed as a group within a work shift,
analyze a single laboratory fortified blank (LFB) containing each
analyte of concern at a concentration as determined in Sect. 9.3.
If more than 20 samples are Included in a batch, analyze a LFB for
every 20 samples.  Use the procedures described in Sect. 9.3.3 to
evaluate the accuracy of the measurements.  If acceptable accuracy
cannot be achieved, the problem must be located and corrected before
additional samples are analyzed.  Add the results to the on-going
control charts to document data quality.

Note: If the LFB for each batch of samples contains the individual
PCB congeners listed in Section 1, then a LFB for each Aroclor is
not required.  At least one LFB containing toxaphene should be
extracted for e.ach 24 hr period during which extractions are
performed.  Toxaphene should be fortified in a separate LFB from
other method analytes.
If individual PCB congeners are not part of the LFB, then it is
suggested that one multi-component analyte (toxaphene, chlordane or
an Aroclor) LFB be analyzed with each sample set.  By selecting a
different multi-component analyte for this LFB each work shift,  LFB
data can be obtained for all of these analytes over the course of
several days.

Determine that the sample matrix does not contain materials that
adversely affect method performance.  This is accomplished by
analyzing replicates of laboratory fortified matrix samples and
ascertaining that the precision, accuracy, and method detection
limits of analytes are in the  same range as obtained with laboratory
fortified blanks.  If a variety of different sample matrices are
analyzed regularly, for example, drinking water from groundwater and
surface water sources, matrix  independence should be established for
each.  Over time, LFM data  should be documented for all routine
sample sources for the laboratory.   A  laboratory fortified sample
matrix should be analyzed for  every 20  samples processed in the same
batch.   If the recovery data for a LFM  does not meet the criteria in
Sect. 9.3.3., and LFBs show the laboratory to be  in control  , then
the  samples from that matrix (sample location) are documented as
suspect due to matrix effects.

With each set of samples, a field reagent blank  (FRB) should be
analyzed.  The results of this analysis will help define
contamination resulting from field sampling and transportation
activities.
                               525.2-16

-------
     9.9  At least quarterly, analyze a quality-control sample from an
          external source.   If measured analyte concentrations are not of
          acceptable accuracy (Sect. 9.3.3), check the entire analytical
          procedure to locate and correct the problem source.

     9.10 Numerous other quality control measures are incorporated into other
          parts of this procedure, and serve to alert the'analyst to
          potential problems.

10.   CALIBRATION AND STANDARDIZATION

     10.1 Demonstration and documentation of acceptable initial  calibration is
          required before any samples are analyzed and is required
          intermittently throughout sample analysis as dictated  by results of
          continuing calibration checks.   After initial  calibration is
          successful,  a continuing calibration  check is required  each  day or
          at  the beginning of each period in which analyses  are  performed not
          to  exceed 12  h.   Additional  periodic  calibration checks are  good
          laboratory practice.   It is recommended  that an  additional
          calibration  check be performed  at  the end of each  period of
          continuous  instrument  operation,  so that all  field sample analyses
          are bracketed by a calibration  check  standard.

     10.2  Initial  calibration

          10.2.1  Calibrate  the mass  and  abundance  scales of  the MS with
                 calibration  compounds and procedures  prescribed  by the
                 manufacturer with  any modifications necessary to  meet the
                 requirements in Sect. 10.2.2.

          10.2.2  Inject  into  the GCf S system a  1 /jl aliquot of the 5 nq/uL
                 solution of  DFTPP, endrin and 4,4'-DDT.  If desired, the
                 endrin and DDT degradation  checks may be performed
                 simultaneously with the DFTPP check or in a separate
                 injection.   Acquire a mass  spectrum that includes data for
                m/z 45-450.  Use GC conditions that produce a narrow (at
                least five scans per peak)  symmetrical peak for each compound
                 (Sect. 10.2.3.1  and Sect. 10.2.3.2).  If the DFTPP mass
                spectrum does not meet all criteria in Table: 1,  the MS must
                be retuned and adjusted to meet all criteria before
                proceeding with  calibration.  A single spectrum or an  average
                spectrum across  the GC peak may be used to evaluate the
                performance of the system.  Locate any degradation products
                of endrin (endrin ketone [EK] and endrin aldehyde FEA1)  and
                4,4'-DDT (4,4'-DDE and 4,4'-DDD).at their appropriate
                retention times  and quantisation ions  (Table 2).   Endrin
                ketone  can be located at =1.1 to 1.2 times the  endrin
                retention time with prominent m/z  67 and 317 ions in the mass
                spectrum.   If degradation of either endrin or DDT exceeds
                20%,  maintenance is required on the GC injection  port  and
                possibly other areas of  the  system before  proceeding with the
                                  525.2-17

-------
       calibration.  Calculate percent breakdown using peak areas
       based on total  ion current (TIC) as follows:
       % 4,4'-DDT breakdown=        ;

         I TIC area of DDT degradation peaks (DDE+DDD)

         I TIC area of total DDT peaks (DDT+DDE+DDD)
X 100
       % endrin breakdown=

         I TIC area of endrin degradation peaks (EA+EK)

         I TIC area of total endrin peaks (endrin+EA+EK)
X 100
10.2.3 Inject a 1-juL aliquot of a medium concentration calibration
       solution, for example 0.5-2 M9/L, and acquire and store data
       from m/z 45-450 with a total cycle time (including scan
       overhead time) of 1.0 sec or less.  Cycle time should be
       adjusted to measure at least :five or more spectra during the
       elution of each GC peak.  Calibration standards for toxaphene
       and Aroclors must be injected individually.

       10.2.3.1 The following are suggested multi-ramp temperature
                program GC conditions.  Adjust the helium carrier
                gas flow rate to about 33 cm/sec.  Inject at 45°C
                and hold in splitless mode for 1 min.  Heat rapidly
                to 130°C.   At 3 min start the temperature program:
                130-180°C at 12°/min;  180-240°C at 7°/min; 240-320°C
                at 12°/min.   Start data acquisition  at 4 min.

       10.2.3.2 Single ramp linear temperature program suggested GC
                conditions.  Adjust the helium carrier gas flow rate
                to about 33 cm/sec.  Inject at 40°C  and hold in
                splitless mode for 1 min.  Heat rapidly to 160°C.
                At 3 min start the temperature program: 160-320°C at
                6°/min;  hold at 320°  for 2  min.   Start data
                acquisition at 3 min.

10.2.4 Performance criteria for the calibration standards.  Examine
       the stored GC/MS data with the data system software.

       10.2.4.1 GC performance.  Anthracene and phenanthrene should
                be separated by baseline.  Benz[a]anthracene and
                chrysene should be separated by a valley whose
                height is less than 25% of the average peak height
                of these two compounds.  If the valley between
                benz[a]anthracene and chrysene exceeds 25%, the GC
                column requires maintenance.  See Sect.  10.3.6.
                                    I
       10.2.4.2 MS sensitivity.  The GC/MS/DS peak identification
                software should be able to recognize  a GC peak in

                         525.2-18

-------
                the appropriate retention time window for each of
                the compounds in the calibration solution, and make
                correct identifications.  If fewer than 99% of the
                compounds are recognized, system maintenance is
                required.  See Sect. 10.3.6.

10.2.5 If all performance criteria are met,  inject a l-/iL aliquot of
       each of the other CAL solutions using the same GC/MS
       conditions.  Calibration standards of toxaphene and Aroclors
       must be injected individually.

       10.2.5.1 Some GC/MS systems may not be sensitive enough to
                detect some of the analytes  in the two lowest
                concentration CAL solutions.  In this case, the
                analyst should prepare additional  CAL solutions at
                slightly higher concentrations to  obtain at least 5
                calibration points that bracket the expected analyte
                concentration range.

10.2.6 Calculate a response factor (RF) for  each analyte of interest
       and surrogate for each CAL solution using the internal
       standard whose retention time is nearest the retention time
       of the analyte or surrogate.   Table 2 contains suggested
       internal  standards for each analyte and surrogate,  and
       quantitation ions for all  compounds.   This  calculation is
       supported in acceptable GC/MS data system software  (Sect.
       6.10.4),  and many other software programs.   The RF  is a
       unitless number,  but units used to express  quantities of
       analyte and internal  standard must be equivalent.

       Note:   To calibrate for multi-component analytes  (toxaphene
       and Aroclors),  one of the  following methods  should  be used.
       Option 1- Calculate an average response factor or linear
       regression  equation for each  multi-component analyte from  the
       combined  area of  all  its component peaks  identified in  the
       calibration standard chromatogram,  using  2-3 of the suggested
       quantitation ions in  Table 2.
       Option 2- Calculate an average response factor or linear
       regression  equation for each  multi-component analyte using
       the combined areas  of 3-6  of  the most  intense  and
       reproducible peaks  in  each  of the calibration  standard
       chromatograms.  Use an appropriate  quantitation ion for  each
       peak.
                     RF =	

                         •CA,.j(Qx)
                        525.2-19

-------
            where:
                     A,   =
                     A  _
                     Mis ~
                     Ql
integrated abundance of the quantitation ion
of the analyte;.
integrated abundance of the quantitation ion
internal  standard.
quantity of analyte injected in ng or
concentration units.
quantity of internal standard injected in ng
or concentration units.
            10.2.6.1 For each analyte and surrogate,  calculate the mean
                     RF from the analyses of the six CAL solutions.
                     Calculate the standard deviation (SD)  and the
                     relative standard deviation (RSD) from each mean:
                     RSD = 100 (SD/M).  If the RSD of any analyte or
                     surrogate mean RF exceeds 30%,  either analyze
                     additional aliquots of appropriate CAL solutions to
                    ' obtain an acceptable RSD of RFs over the entire
                     concentration range, or take action to improve GC/MS
                     performance.  See Sect. 10.3.6.

     10.2.7 As an alternative to calculating mean response factors, use
            the GC/MS data system software or other available software to
            generate a linear regression calibration by plotting Ax /Ais
            vs. Qx.

10.3 Continuing calibration check.  Verify the MS tune and initial
     calibration at the beginning of each 12 h work shift during which
     analyses are performed using the following procedure.

     10.3.1 Inject a 1-0L aliquot of the |5 ng//uL solution of DFTPP,
            endrin, and 4,4'-DDT.  Acquire a mass spectrum for DFTPP that
            includes data for m/z 45-450.  Ensure that all criteria in
            Sect. 10.2.2 are met.
     10.3.2 Inject a l-/iL aliquot of a calibration solution
            with the same conditions used during the initial
            It is recommended that the concentration of cali
            solution be varied, so that the calibration can
            at more than one point.  Note: If the continuing
            check standard contains the PCB congeners listed
            1, calibration verification js not required for
            Calibration verification of toxaphene should be
            least once each 24 hr period.
                                 and analyze
                                  calibration.
                                 bration
                                 be verified
                                  calibration
                                  in Section
                                 each Aroclor.
                                 performed  at
     10.3.3 Demonstrate acceptable performance for the criteria  shown  in
            Sect.  10.2.4.

     10.3.4 Determine that  the  absolute  areas of the quantitation  ions of
            the  internal  standards and surrogate(s) have not changed by
            more than 30% from  the areas measured in the most  recent
            continuing calibration check;  or by more than  50%  from the
                               525.2-20

-------
       areas measured during initial calibration.  If these areas
       have decreased by more than these amounts, adjustments must
       be made to restore system sensitivity.   These adjustments may
       require cleaning of the MS ion source,  or other maintenance
       as indicated in Sect. 10.3.6,,and recalibration.   Control
       charts are useful aids in documenting system sensitivity
       changes.

10.3.5 Calculate the RF for each analyte and surrogate from the data
       measured in the continuing calibration  check.  The RF for
       each analyte and surrogate must be within 30% of the mean
       value measured in the initial calibration.  Alternatively, if
       a linear regression is used,  the calculated amount for each
       analyte must be ± 30% of the  true value.   If these conditions
       do not exist, remedial action should be taken which may
       require recalibration.  Any field sample  extracts that have
       been analyzed since the last  acceptable calibration
       verification should be reanalyzed after adequate calibration
       has been restored.

       10.3.5.1 Because of the large number of compounds on the
                analyte list, it is  possible for a few analytes.of
                interest to.be outside the continuing calibration
                criteria.  If analytes that missed the calibration
                check are detected in samples,  they may be
                quantified using a single point  calibration.  The
                single point standards should  be prepared at
                concentrations that  produce responses close (±20%)
                to those of the unknowns.  If  the same analyte
                misses the continuing calibration check on three
                consecutive work shifts, remedial action MUST be
                taken.  If more than 10% of the  analytes of interest
                miss the continuing  calibration  check on a single
                day, remedial action MUST.be taken.

10.3.6 Some possible remedial actions.  Major  maintenance such as
       cleaning an ion source, cleaning quadrupole rods, replacing
       filament assemblies, etc. require returning to the initial
       calibration step.

       10.3.6.1 Check and adjust GC  and/or MS  operating conditions;
                check the MS resolution, and calibrate the mass
                scale.

       10.3.6.2 Clean or replace the splitless injection liner;
                silanize a new injection liner.

       10.3.6.3 Flush the GC column  with solvent according to
                manufacturer's instructions.

       10.3.6.4 Break off a short portion (about 1 meter) of the
                column from the end  near the injector; or replace GC

                         525.2-21

-------
                          column.  This action will cause a change in
                          retention times.

                 10.3.6.5 Prepare fresh CAL ^solutions, and repeat the initial
                          calibration step.

                 10.3.6.6 Clean the MS ion source and rods (if a quadrupole).
                                            j:
                 10.3.6.7 Replace any components that allow analytes to come
                          into contact with hot metal surfaces.

                 10.3.6.8 Replace the MS electron multiplier, or any other
                          faulty components.
11.   PROCEDURE
     11.1  CARTRIDGE EXTRACTION

          11.1.1 This procedure may be carried out in the manual mode or in
                 the automated mode (Sect. 6.12) using a robotic or automatic
                 sample preparation device.  If an automatic system is used to
                 prepare samples, follow the manufacturer's operating
                 instructions, but follow this procedure.  If the manual mode
                 is used, a suggested setup of the extraction apparatus is
                 shown in Figure 1A.  The reservoir is not required, but
                 recommended for convenient operation.  Water drains from the
                 reservoir through the LSE cartridge and into a syringe needle
                 which is inserted through a rubber stopper into the suction
                 flask.  A slight vacuum of approximately 13 cm (5 in.) of
                 mercury is used during all operations with the apparatus.
                 About 2 h should be required to draw a liter of water through
                 the cartridge.

          11.1.2 Elute each cartridge with a 5 mL aliquot of ethyl acetate
                 followed by a 5 mL aliquot of methylene chloride.  Let the
                 cartridge drain dry after each flush.  Then elute the
                 cartridge with a 10 mL aliquot of methanol, but DO NOT allow
                 the methanol  to elute below the top of the cartridge packing.
                 From this point, do not allow the cartridge to go dry.  Add
                 10 mL of reagent water to the cartridge, but before the
                 reagent water level drops below the top edge of the packing,
                 begin adding  sample to the solvent reservoir.

          11.1.3 Pour the water sample into the 2-L separatory funnel with the
                 stopcock closed, add 5 mL methanol, and mix well.  If a
                 vacuum manifold is used instead of the separatory funnel, the
                 sample may be transferred directly to the cartridge after the
                 methanol is added to the sample. (Residual chlorine should
                 not be present as a reducing agent should have been added at
                 the time of sampling.  Also the pH of the sample should be
                 about 2.  If  residual chlorine is present and/or the pH is
                 >2, the sample may be invalid.)  Add a 100-/JL aliquot of the

                                   525.2-22

-------
            fortification solution (50 /Kj/mL) for internal standards and
            surrogates, and mix immediately until homogeneous.  The
            resulting concentration of these compounds in the water
            should be 5
     11.1.4 Periodically transfer a portion of the sample into the
            solvent reservoir.  The water sample will drain into the
            cartridge, and from the exit into the suction flask.
            Maintain the packing material in the cartridge immersed in
            water at all times.  After all  of the sample has passed
            through the LSE cartridge, draw air or nitrogen through the
            cartridge for 10 min.

     11.1.5 Transfer the 125-mL solvent reservoir and LSE cartridge (from
            Figure 1A) to the elution apparatus if used (Figure IB).  The
            same 125-mL solvent reservoir is used for both apparatus.
            Rinse the inside of the 2-L separatory funnel  and the sample
            jar with 5 mL of ethyl acetate, and elute the cartridge with
            this rinse into the collection tube.  Wash the inside of the
            separatory funnel and the sample jar with 5 mL methylene
            chloride and elute the cartridge, collecting the rinse in  the
            same collection tube.  Small  amounts of residual  water from
            the sample container and the LSE cartridge may form an
            immiscible layer with the eluate.  Pass the eluate through
            the drying column (Sect. 6.7) which is packed with
            approximately 5 to 7 grams of anhydrous sodium sulfate and
            collect in a second vial.  Wash the sodium sulfate with at
            least 2 mL methylene chloride and collect in the same vial.
            Concentrate the extract in a warm water bath under a gentle
            stream of nitrogen.   Do not concentrate the extract to less
            than 0.5 mL, as this will result in losses of analytes.  Make
            any volume adjustments with ethyl acetate.  It is recommended
            that an aliquot of the recovery standard be added to the
            concentrated extract to check the recovery of the internal
            standards (see Sect. 7.12).

11.2  DISK EXTRACTION

     11.2.1 This procedure was developed using the standard 47 mm
            diameter disks.   Larger disks (90 mm diameter) may be used if
            sample compositing is being done or special  matrix problems
            are encountered.   If larger disks are used,  the washing
            solvent volume is 15 mL, the conditioning solvent volume is
            15 mL,  and the elution solvent  volume is two 15 mL aliquots.

            11.2.1.1 Extractions using the  disks may be carried out
                     either in the manual  or automatic mode (Sect.  6.12)
                     using an automatic sample preparation device.   If an
                     automatic system is  used to prepare samples,  follow
                     the manufacturer's operating instructions,  but
                     follow this procedure.   Insert the disk into the
                     filter apparatus (Figure 2)  or sample preparation

                              525.2-23

-------
                unit.  Wash the disk with 5 ml of a 1:1 mixture of
                ethyl acetate (EtAc) and methylene chloride (MeC12)
                by adding the solvent to the disk, drawing about
                half through the disk, allowing it to soak the disk
                for about a minute, then drawing the remaining
                solvent through the disk.  (NOTE:  Soaking the disk
                may not be desirable when disks other than Teflon
                are used.  Instead, apply a constant, low vacuum in
                this Section and Sect. 11.2.1.2 to ensure adequate
                contact time between solvent and disk.)

       11.2.1.2 Pre-wet the disk with 5 ml methanol (MeOH) by adding
                the MeOH to the disk and allowing it to soak for
                about a minute, then drawing most of the remaining
                MeOH through.  A layer of MeOH must be left on the
                surface of the disk, which should not be allowed to
                go dry from this point until the end of the sample
                extraction.  THIS IS A CRITICAL STEP FOR A UNIFORM
                FLOW AND GOOD RECOVERY.

       11.2.1.3 Rinse the disk with 5 mL reagent water by adding the
                water to the disk and; drawing most through, again
                leaving a layer on the surface of the disk.

11.2.2 Add 5 mL MeOH per liter of water to the sample.  Mix well.
       (Residual chlorine should not be present as a reducing agent
       should have been added at the time of sampling.  Also the pH
       of the sample should be about 2.  If residual chlorine is
       present and/or the pH is >2, the sample may be invalid.)

11.2.3 Add 100 ill of the internal standard and surrogate compound
       fortification solution (50 /ig/mL) to the sample and shake or
       mix until the sample is homogeneous.  The resulting
       concentration of these compounds in the water should be 5
11.2.4 Add the water sample to the reservoir and apply full vacuum
       to begin the extraction.  Particulate-free water may pass
       through the disk in as little as 5 min without reducing
       analyte recoveries.  Extract the entire sample, draining as
       much water from the sample container as possible.  Dry the
       disk by maintaining vacuum for about 10 min.

11.2.5 Remove the filtration top, but do not disassemble the
       reservoir and fritted base.  If a suction flask is being
       used, empty the water from the flask, and insert a suitable
       collection tube to contain the eluant.  The only constraint
       on the sample tube is that it fit around the drip tip' of the
       fritted base.  Reassemble the apparatus.

11.2.6 Add 5 mL of ethyl acetate to the sample bottle, and rinse the
       inside walls thoroughly.  Allow the solvent to settle to the
       bottom of the bottle, then transfer it to the disk.  A
       disposable pipet or syringe may be used to do this, rinsing
       the sides of the glass filtration reservoir in the process.

                         525.2-24

-------
            Draw about half of the solvent through the disk, release the
            vacuum, and allow the disk to soak for a minute.  Draw the
            remaining solvent through the disk.  (NOTE:  Soaking the disk
            may not be desirable if disks other than Teflon are used.
            Instead, apply a constant, low vacuum in this Section and
            Sect. 11.2.7 to ensure adequate contact time between solvent
            and disk.)

     11.2.7 Repeat the above step (Sect. 11.2.6) .with methylene chloride.

     11.2.8 Using a syringe or disposable pipet, rinse the filtration
            reservoir with two 3 ml portions of 1:1 EtAc:MeC12.  Draw the
            solvent through the disk and into the collector tube.  Pour
            the combined eluates (Sect. 11.2.6, Sect. 11.2.7, and Sect.
            11.2.8) through the drying tube (Sect.  6.7) containing about
            5 to 7 grams of anhydrous sodium sulfate.  Rinse the drying
            tube and sodium sulfate with two 3 ml portions of 1:1
            EtAc:MeC12 mixture.  Collect all the extract and washings in
            a concentrator tube.

     11.2.9 While gently heating the extract in a water bath or a heating
            block, concentrate to between 0.5 and 1 mL under a gentle
            stream of nitrogen.  Do not concentrate the extract to less
            than 0.5 mL, since this will result in  losses.of analytes.
            Make any volume adjustments with ethyl  acetate.  It is  '
            recommended that an aliquot of the recovery standard be added
            to the concentrated extract to check the recovery of the
            internal standards (see Sect. 7.12).

11.3 Analyze a 1 /iL aliquot with the GC/MS system under the same
     conditions used for the initial  and continuing calibrations (Sect.
     10.2.3).

11.4 At the conclusion of data acquisition,  use the same software that
     was used in the calibration procedure to tentatively identify peaks
     .in predetermined retention time windows of interest.  Use the data
     system software to examine the ion abundances  of components of the
     chromatogram.

11.5 Identification of analytes.  Identify a.sample component by
     comparison of its mass spectrum (after background subtraction)  to a
     reference spectrum in the user-created data base.  The GC retention
     time of the sample component should be within  5 sec of the retention
     time observed for that same compound in the most recently analyzed
     continuing calibration check standard.

     11.5.1 In general,  all ions that are present above 10% relative
            abundance in the mass spectrum of the standard should be
            present in the mass spectrum of the sample component and
            should agree within absolute 20%.   For  example, if an ion has
            a relative abundance of 30% in the standard spectrum.,  its
            abundance in the sample spectrum should be in the range  of 10

                              525.2-25

-------
                 to 50%.  Some  ions, particularly the molecular ion, are of
                 special  importance, and should be evaluated even if they are
                 below 10% relative abundance.

          11.5.2 Identification  is hampered when sample components are not
                 resolved chromatographically and produce mass spectra
                 containing ions contributed by more than one analyte.  When
                 GC peaks obviously represent more than one sample component
                 (i.e., broadened peak with shoulder(s) or valley between two
                 or more maxima), appropriate analyte spectra and background
                 spectra can be  selected by examining plots of characteristic
                 ions for tentatively identified components.  When analytes
                 coelute  (i.e.,  only one GC peak is apparent), the
                 identification  criteria can be met but each analyte spectrum
                 will contain extraneous ions contributed by the coeluting
                 compound.

          11.5.3 Structural isomers that produce very similar mass spectra can
                 be explicitly  identified only if they have sufficiently
                 different GC retention times.  See Sect. 10.2.4.1.
                 Acceptable resolution is achieved if the height of the valley
                 between two isomer peaks is less than 25% of the average
                 height of the two peak heights.  Otherwise, structural
                 isomers are identified as iisomeric pairs.  Benzo[b] and
                 benzo[k]fluoranthene may be measured as an isomeric pair.
                 MGK 264 is made up of two structural isomers.  These are
                 listed separately in the data tables.
                                           i
          11.5.4 Each multi-component analyse can be identified by the
                 presence of its individual components in a characteristic
                 pattern based on the relative amounts of each component
                 present.  Chromatograms of standard materials of multi-
                 component analytes should be carefully evaluated, so that
                 these patterns can be recognized by the analyst.
                                           1
12.  DATA ANALYSIS AND CALCULATIONS        ;

     12.1 Complete chromatographic resolution is not necessary for accurate
          and precise measurements of analyte concentrations if unique ions
          with adequate intensities are avail able for quantitation.   In
          validating this method, concentrations were calculated by measuring
          the characteristic ions listed in Table 2.   If the response of any
          analyte exceeds the calibration riage established in Section 10,
          dilute the extract and reanalyze.

          12.1.1 Calculate analyte and surrogate concentrations,  using the
                 multipoint calibration established in Sect. 10.   Do not use
                 daily calibration verification data to quantitate analytes in
                 samples.                  ;
                                  (Ais) RF V

                                   525.2-26

-------
                 where:
                     A.
                    As =

                     Qis -
                     v   =
                     RF  =
concentration of analyte or surrogate in (j.g/1 in
the water sample.
integrated abundance of the quantitation ion of
the analyte in the sample.
integrated abundance of the quantitation ion of
the internal  standard in the sample.
total  quantity (in micrograms) of internal
standard added to the water sample.
original water sample volume in liters.
mean, response factor of analyte from the initial
calibration.   RF is a unitless value.
          12.1.2 Alternatively,  use the GC/MS system software or other
                 available proven software to compute the concentrations  of
                 the analytes and surrogates from the linear regression
                 established in  Sect.  10.   Do not use daily calibration
                 verification data to  quantitate analytes in samples.

          12.1.3 Calculations should utilize all available digits of
                 precision,  but  final  reported concentrations should be
                 rounded to an appropriate number of significant figures  (one
                 digit of uncertainty).  Experience indicates that three
                 significant figures may be used for concentrations above 99
                       two significant figures for concentrations between 1-99
                       and one significant figure for lower concentrations.
     12.2 To quantitate multi-component analytes (toxaphene and Aroclors),  one
          of the following methods should be used.

          Option 1 - Calculate an average RF or linear regression equation  for
          each multi -component analyte from the combined area of all  its
          component peaks identified in the calibration standard chromatogram,
          using 2-3 of the suggested quantitation ions in Table 2.

          Option 2 - Calculate an average response  factor or linear regression
          equation for each multi-component analyte using the combined areas
          of 3-6 of the most intense and reproducible peaks in each of the
          calibration standard chromatograms.

          When quantifying multi-component analytes in samples, the analyst
          should use caution to include only those  peaks from the sample that
          are attributable to the multi-component analyte.  Option 1  should
          not be used if there are significant interference peaks within the
          Aroclor or toxaphene pattern.  Option 2 was used to generate the
          data in Table 6.

13.   METHOD PERFORMANCE

     13.1 Single laboratory accuracy and precision  data (Tables 3-6)  for each
          listed analyte (except multi-component analytes) were obtained at a

                                   525.2-27

-------
     concentration of 0.5 /*g/L and/or 5 fJy/L in reagent water utilizing
     both the disk and the cartridge technology and two different GC/MS
     systems, an ion trap and a quadrupole mass spectrometer.  Table 8
     lists accuracy and precision data from replicate determinations of
     method analytes in tap water using liquid-solid cartridge
     extractions and the ion trap mass spectrometer.  Any type of GC/MS
     system may be used to perform this method if it meets the
     requirement in Sect. 6.10 and the quality control criteria in Sect.
     9.  The multi-component analytes (i.e. toxaphene and Aroclors) are
     presented in Tables 5 and 6.  The average recoveries in the tables
     represent six to eight replicate analyses done over a minimum of a 2
     day period.

     13.1.2 With these data, the method detection limits (MDL) in the
            tables were calculated using the formula:
MDL = S t^..,

where:

t,,, 1 1 „!„.,, _ n oo>
 \T\" I • \ "aipna — u«yy j
                                0-99)
                            =  Student's  t  value  for  the  99%  confidence
                                        _ ,  ,  .       .        r- r-    \
                                    level with n-1 degrees of freedom
            n = number of replicates
            S - standard deviation of replicate analyses.

13.2 Problem compounds

     13.2.1 Some polycyclic aromatic hydrocarbons (PAH), including the
            labeled PAHs used in this method as internal standards, are
            rapidly oxidized and/or chlorinated in water containing
            residual chlorine.  Therefore, residual chlorine must be
            reduced at the time of sampling.   .These same types of
            compounds, especially anthracene, benz[a]anthracene, and
            benzo[a]pyrene, are susceptible to photodegradation.
            Therefore, care should be taken to avoid exposing standards,
            samples, and extracts to direct light.  Low recoveries of
            some PAH compounds have been observed when the cartridge or
            disk was air dried longer than 10 min (Sect. 11.1.4 and Sect.
            11.2.4).  Drying times longer than 10 min should be avoided,
            or nitrogen may be used to dry the cartridge or disk to
            minimize the possible oxidation of these analytes during the
            drying step.

     13.2.2 Merphos is partially converted to DEF in aqueous matrices,
            and also when introduced into a hot gas chromatographic
            injection system.  The efficiency of this conversion appears
            to be unpredictable and not reproducible.  Therefore, merphos
            cannot be quantified and can only be identified by the
            presence of DEF in the sample.

     13.2.3 Several of the nitrogen and/or phosphorus containing
            pesticides listed as method analytes are difficult to

                              525.2-28

-------
         chromatograph  and  appear as  broad,  asymmetrical  peaks.  These
         analytes,  whose  peak shapes  are  typically  poor,  are  listed  in
         Table  7.   The  method performance for  these analytes  is
         strongly dependent  on chromatographic efficiency and
         performance.   Poor  peak  shapes will affect the  linearity  of
         the  calibration  curves and result  in  poor  accuracy at low
         concentrations.  Also listed  in  Table 7  are data generated  at
         a mid-concentration  level for these analytes.   In most  cases,
         the  data at this concentration meet the  quality  control
         criteria requirements of the method.

 13.2.4  Phthalate  esters and  other background components  appear in
         variable quantities  in laboratory and field reagent  blanks,
         and  generally  cannot  be  accurately measured at levels below
         about  2 /tg/L.  Subtraction of the concentration  in the blank
         from the concentration in the sample  at  or  below  the 2 jag/L
         level  is not recommended  because the  concentration of the
         background in  the blank  is highly variable.

 13.2.5 Atraton and prometon  are  not efficiently extracted from the
        water  at pH 2  due to  what appears to  be  their ionization
        occurring  in solution under acidic conditions.   In order to
        determine these analytes  accurately,  a separate sample must
        be collected and dechlorinated with sodium sulfite,  but no
        HC1  should be  added at.the time of collection.  At neutral
        pH,  these two  compounds are recovered  from water with
        efficiencies greater than 90%.  The data in Tables 3, 4, 5,
        6,  and 8 are from samples extracted at pH 2.

 13.2.6 Carboxin,  disulfoton, and disulfoton sulfoxide were  found  to
        be  unstable in  water and  began to degrade almost immediately.
        These analytes  may be identified  by this method but  not
        accurately  measured.

 13.2.7 Low  recoveries  of metribuzin  were observed  in  samples
        fortified with  relatively high concentrations  of additional
        method  analytes.   In samples  fortified with approximately  80
        analytes at 5 /ig/L  each,  metribuzin  was recovered at  about
        50%  efficiency.  This suggests  that  metribuzin may break
        through the C-18  phase in highly  contaminated  samples
        resulting  in  low  recoveries.

"13.2.8 If cyanazine  is to  be determined,  a  separate sample must be
        collected.  Cyanazine degrades  in the  sample when it  is
        stored  under acidic  conditions or when sodium  sulfite is
        present in  the  stored sample.  Samples collected  for
        cyanazine determination MUST NOT  be dechlorinated or
        acidified when  collected.  They should be iced or
        refrigerated and  analyzed  within  14 days.   However, these
        samples MUST be dechlorinated and  acidified  immediately prior
        to fortification with  internal standards  and surrogates, and
        extraction  using the  same  quantities of acid and  sodium
        sulfite described in  Sect. 8.

                         525.2-29

-------
14.  POLLUTION PREVENTION

     14.1 This method utilizes liquid-solid extraction (LSE) technology to
          remove the analytes from water.  It requires the use of very small
          volumes of organic solvent and very small quantities of pure
          analytes, thereby eliminating the potential hazards to both the
          analyst and the environment involved with the use of large volumes
          of organic solvents in conventional liquid-liquid extractions.
                                            I
     14.2 For information about pollution prevention that may be applicable to
          laboratory operations, consult "Less Is Better:  Laboratory Chemical
          Management for Waste Reduction" available from the American Chemical
          Society's Department of Government Relations and Science Policy,
          1155 16th Street N.W., Washington, D.C., 20036.

15.  WASTE MANAGEMENT

     15.1 It is the laboratory's responsibility to comply with all federal,
          state, and local regulations governing waste management, particu-
          larly the hazardous waste identification rules and land disposal
          restrictions.  The laboratory using this method has the respons-
          ibility to protect the air, water, and land by minimizing and
          controlling all releases from fume hoods and bench operations.
          Compliance is also required with any sewage discharge permits and
          regulations.  For further information on waste management, see "The
          Waste Management Manual for Laboratory Personnel," also avail-able
          from the American Chemical Society at the address in Sect. 14.2.

16.  REFERENCES

1.   Glaser, J. A., D. L. Foerst, G. D. McKee, S. A. Quave, and W. L. Budde,
     "Trace Analyses for Wastewaters," Environ. Sci. Technol. 1981 15.,
     1426-1435. or 40 CFR, Part 136, Appendix B.

2.   "Carcinogens - Working With Carcinogens," Department of Health,
     Education, and Welfare, Public Health Service, Center for Disease
     Control, National Institute for Occupational Safety and Health,
     Publication No. 77-206, Aug. 1977.

3.   "OSHA Safety and Health Standards, General Industry," (29CFR1910),
     Occupational Safety and Health Administration, OSHA 2206, (Revised,
     January 1976).

4.   "Safety in Academic Chemistry Laboratories," American Chemical Society
     Publication, Committee on Chemical Safety, 3rd Edition, 1979.

5.   Junk, G. A., M. J. Avery, J. J. Richard, "Interferences in Solid-Phase
     Extraction Using C-18 Bonded Porous Silica Cartridges," Anal. Chem. 1988,
     60, 1347.
                                   525.2-30

-------
 17.   TABLES.  DIAGRAMS.  FLOWCHARTS.  AND VALIDATION DAtA
        TABLE 1.   ION ABUNDANCE CRITERIA FOR BIS(PERFLUOROPHENYL)PHENYL
                 •  PHOSPHINE (DECAFLUOROTRIPHENYLPHOSPHINE,  DFTPP)
 Mass     Relative Abundance
 (M/z)	Criteria	
 Purpose of Checkpoint1
  51     10-80% of the  base  peak

  68     <2%  of mass  69

  70     <2%  of mass  69

 127     10-80% of the  base  peak

 197     <2%  of mass  198

 198     base peak or >50% of  44'2

 199     5-9% of mass 198

 275     10-60% of the  base  peak

 365     >1%  of the base peak

 441     Present and  <  mass  443

 442     base peak or >50% of  198

 443     15-24%  of mass 442
.low mass sensitivity

 low mass resolution

 low mass resolution  '

 low-mid  mass  sensitivity

 mid-mass resolution

 mid-mass resolution  and sensitivity

 mid-mass resolution  and.isotope ratio

 mid-high mass  sensitivity

 baseline threshold

 high mass resolution

 high mass resolution: and sensitivity

 high mass resolution and isotope ratio
 All  ions are used primarily to check the mass
spectrometer and data system, and this is the
performance test.  The three resolution checks
abundance isotope ratios, constitute the next'
performance test.  The correct setting of the
by the presence of low intensity ions, is the
performance test.  Finally, the ion abundance
some standardization to fragmentation patterns
          measuring accuracy of the mass
         most important part of the
         , which include natural
         most important part of the
         baseline threshold, as indicated
         next most important part of the
         ranges are designed to encourage
                                   525.2-31

-------
TABLE 2.
RETENTION TIME DATA, QUANTITATION IONS, AND INTERNAL STANDARD REFERENCES FOR METHOD ANALYTES
"
Compound
Retention
Time (min:sec)
i A" B"
Quant i tat ion
Ion
IS
Reference
#
'
Internal Standards
acenaphthene-dIO (#1)
chrysene-d12 (#2)
phenanthrene-d10 (#3)
7:47
21 :33
11:37
7:01
18:09
10:13
164
240
188



.
Surrogates
1 ,3-dimethyl-2-nitrobenzene
perylene-d12
triphenylphosphate
5:16
26:60
20:25
4:33
21:31
17:25
134
264
326/325
1
3
3
' . !
Target Analytes
acenaphthylene
alachlor
aldrin
atnetryn
anthracene
Aroclor 1016
ArocLor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
atraton
atrazine
benz [a] anthracene
benzo [blfluoranthene
benzo [k] fluoranthene
benzo [g,h,i]perylene
benzo [a] pyrene
bromaci I
butachlor
butylate
butylbenzylphthalate
carboxin
7:30
12:59
14:24
13:11
f11:50






,-,
10:31
10:49
21 :31
25:33
25:45
31:16
25:24
13:46
16:25
I 6:60
19:39
17:37
6:46
11:24
12:31
11:35
10:24
7:30-14:00
6:38-11:25
6:38-13:54
6:38-15:00
8:47-15:00
11:00-18:00
13:10-21:00
9:25
9:38
18:08
20:44
20:48
24:18
21:25
12:03
14:16
6:23
16:53
15:13
152
160
66
227/170
178
152/256/292
152/222/256
152/256/292
152/256/292
152/256/292
220/326/360
326/360/394
196/169
200/215
228
252
252
276
252
205
176/160
57/146
149
143
1
2
2
2
2
2
2
2
2
2
2
2
1
1/2
3
3
3
3
3
2
2
1
2/3
2
                                                  525.2-32

-------
TABLE 2.   RETENTION TIME DATA, QUANTITATION IONS,  AND INTERNAL STANDARD REFERENCES FOR METHOD ANALYTES
          (CONTINUED)
• • • • -• II
Compound
Retention
..... Time (min:sec)
Aa Bb
Quant i tat ion
Ion

chlordane, (alpha-chlordane)
chlordane, (gamma-chlordane)
chlordane, (trans-nonachlor)
chlorneb
chlorobenzilate
2-chlorobiphenyl ,
chlorprophara
chlorpyrifos
chlorothalonil
chrysene
cyanazine
cycloate
DCPA
4,4'-DDD
4,4'-DDE
4,4'-DDT
DEF
diazinon
dibenz [ [a, h] anthracene
di-n-butylphthalate
2,3-dichlorobiphenyl
dichlorvos
dieldrin
di(2-ethylhexyl)adfpate
di(2-ethylhexyl)phthalate
diethylphthalate
dimethylphthalate
2,4-dinitrotoluene . . ,
2,6-dimtrotoluene
diphenamid
disulfoton
disulfoton sulfone
disulfoton sulf oxide
16:43
16:19
16:47
7:47
18:22
7:53
9:33
14:10
11:38
21:39
14:14
9:23
14:20
18:40
17:20
19:52
17:24
11:19
30:32
13:49
10:20
5:31
17:35
20:11
22:11
8:68
7:13
8:08
7:19
14:52
11:43
16:28
6:09
14:28
14:05
14:30
7:05
15:52
7:08
8:36
12:23
10:15
18:13
12:28
8:26
12:30
16:05
14:59
17:00
15:05
10:05
23:47
12:07
9-12
4:52
15:09
17:19
18:39
7:53
6:34
7-22
6:40
12:58
10:22
14:17
	 5:31
375/373
373
409
191
139
188
127
197/97
266
228
225/68
83/154
301
235/165
246
235/165
57 /169
137/179
278
149

109
79

149
149
163
165
165


213/153

IS
Reference
#


2/3
2/3
1
2
1
1



2
1


2



3

1
1



1
1
1
1



1
                                                525.2-33

-------
TABLE 2.  RETENTION TIME DATA, QUANTITATION IONS, AND INTERNAL STANDARD REFERENCES FOR METHOD ANALYTES
          (CONTINUED)                                     ;

Compound
Retention
Time (minisec)
Aa B"

endosulfan I
endosulfan II
endosulfan sulfate
endrin
endrin aldehyde
EPTC
ethoprop
etridiazole
fenamiphos
fenarimol
fluorene
fluridone
HCH, alpha 	
HCH, beta
HCH, delta
HCH, gamma (Lindane)

heptachlor epoxide
2,2' ,3,3' ,4,4' ,6-heptachlorobiphenyl
hexach lorobenzene
2,2',4,4',5,6'-hexachlorobiphenyl

hexazinone
indenoC1r2,3-cd]pyrene
isophorone
merphos
methoxychlor
methyl paraoxon
tnetolachlor
metribuzin
mevinphos
HCK 264 - isomer a
MGK 264 - isomer b
16:44
18:35
19:47
18:15
19:02
6:23
9:19
7:14
16:48
23:26
8:59
26:51
10:19
10:57
11:57
11:13
13:19
15:34
21:23
10:27
17:32
5:16
20:00
30:26
4:54
15:38
21 :36
11:57
14:07
12:46
5:54
15:18
14:55
14:26
15:59
.' 16:54
15:42
16:20
5:46
8:23
6:37
14:34
19:24
8:03
21:26
. 9:10
9:41
10:32
9:54
11:37
13:29
18:04
9:15
15:09
5:38
17:06
23:43
4:10
13:35
18:14
10:22
12:20
11:13
6:19
13:00
13:19

Quant i tat ion
Ion

195
195
272
67/81
67
128
158
211/183
303/154
139
166
328
181
181
181
181
100
81
394/396
284
360
237
171
276
82
209/153
227
109
162
198
127
164/66
164

IS
Reference
#

' 2
2
2
2
2
1
1
1
2
3
1
-3 ••"'
1
2
2
2
2
• 2 :
3
1
2
1
2
3
1
2
3
2
2-
2
1
2
2
                                                   525.2-34

-------
TABLE 2.
                         DATA, QUANTITATION IONS. AND INTERNAL STANDARD REFERENCES FOR METHOD ANALYTES
• - — 	 II
Compound
Retention
Time (min:sec)
Aa B"

molinate
napropamide
norflurazon .
2,2' ,3,3' ,4,5' ,6,6'-octachlorobiphenyl
pebulate
2,2',3',4,6-pentachlorobiphenyl
pentachlorophenol
permethrin, cis
permethrin, trans
phenanthrene
prometon
prometryn
pronamide
propachlor
propazine
pyrene
simazine
simetryn
stirofos
tebuthiuron
terbacil
terbufos
terbutryn
2,2',4,4'-tetrachlorobiphenyl
toxaphene
8:19
16:53
19:31
21 '33
7:18
15:37
11:01
24:25
24:39
11:41
10:39
13:15
11:19
9:00
10:54
16:41
10:41
13:04
16:20
8:00
11:44
11:14
13:39
14:02

triademefon I 14-30
2,4,5-trichlorobiphenyl
tricyclazole
trifluralin
vernolate
Single-ramp linear temperature program conditions
Multi-ramp linear temperature program conditions (J
12:44
17:15
9:31
7:30
14:37
16:46
18:11
6:40
13:33
9:45
20:01
20:10
10:16
9:32
. 11:39
10:02
8:07
9:43

9:33
11:29
14:11
7:16
10:24
9:58
11:58
12:14
13:00-21:00
12:40
10:53
14:51
8:37
7:10 I 	 6:32
[Sect. 10.2.3.2).
Sect. 10.2.3.1).
Quant i tat ion
Ion

126

145
430/428
128


183
183
178
225/168
241/184
173
120
214/172

201/186
213
109
156
161

226/185
292
159
IS
Reference
#



_



2
3
	
3

2
,
2
1
,

2

1
2

2
2-
?
57

189
306
128
. . '

2 II
~^H
-^— I
                                               525.2-35

-------
TABLE 3          ACCURACY AND PRECISION DATA FROM EIGHT DETERMINATIONS OF THE METHOD ANALYTES IN  REAGENT WATER
                 USING LIQUID-SOLID C-18 CARTRIDGE EXTRACTION AND THE QUADRUPOLE MASS SPECTROMETER


Compound
True
Cone.
C/ig/U
Mean
Observed
Cone .
(;ig/L)
Relative
Standard
Deviation
(%)

Mean Method
Accuracy
(% of True
Cone.)

MDL
(/ig/D
[.
Surrogates 	 r
1,3-dimethyl-2-m'trobenzene
ucrylene-d12
triphenylphosphate
5.0
5.0
5.0
4.7
4.9
'• • 5.5
3.9
4.8
6.3

94
98
110





Target Analytes
acenaphthylene
alachlor
aldrin
ametryn
anthracene
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroctor 1242
Aroclor 1448
Aroclor 1254
Aroclor 1260
atraton4
atrazine
bcnz [a] anthracene
benzo [b] f luoranthene

benzo[g.h,i]perylene
bcnzo [a] pyrene
bromacil
butachlor
butylate
butylbenzylphthalate
carboxin
chlordane Calpha-chlordane)
chlordane (ganroa-chlordane)

0.50 :
0.50
0.50
0.50 '
0.50
ND-
ND
ND:
ND,
ND
ND;
ND
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
. 5.0
0.50
0.50
0.50
0.45
:o.47
0.40
: 0..44
0.53
' ND
ND
ND
• ND
ND
ND
. ND
0.35
0.54
0.41
0.49
0.51
0.72
0.58
0.54
0.62
0.52
0.77
3.8
0.36
0.40
0.43

8.2
12
9.3
6.9
' 4.3
ND
ND
ND
ND
ND
ND
ND
15
4.8
16
20
35
2.2
1.9
6.4
4.1
4.1
11
12
11
8.8
17

91
93
80
88
106
ND
ND
ND
ND
ND
ND
ND
70
109
82
98
102
144
116
108
124
105
154
76
72
80
87

0.11
0.16
0.11
0.092
0.068
ND
ND
ND
ND
ND
ND
ND
0.16
0.078
0.20
0.30
0.54
0.047
0.032
0.10
0.076
0.064
0.25
1.4
0.12
0.11
0.22' I
                                                      525.2-36

-------
TABLE 3.         ACCURACY AND PRECISION DATA FROM EIGHT DETERMINATIONS OF  THE  METHOD  ANALYTES IN REAGENT WATER USING
                 LIQUID-SOLID C-18 CARTRIDGE  EXTRACTION AND  THE QUADRUPOLE  MASS SPECTROMETER
                 (CONTINUED)                     :          :


Compound
True ;
Cone .
(/ig/L)
Mean
Observed
Cone.
(Jig/L)
Relative
Standard
Deviation
(%)
Mean Method
Accuracy
(% of True
Cone.)

chlorneb •
chlorobenzilate
2-chlorobiphenyl
chlorpropham
chlorpyrifos • :
chlorothaloni t
chrysene
cyanazine
cycloate
DCPA
4,4'-DDD
4,4'-DDE
4,4'-DDT •
diazinon
dibenz [a, h] anthracene
di-n-butylphthalate
2,3-dichlorobiphenyl
dichtorvos
dieldrin
di-(2-ethylhexyl)adipate
di'(2-ethylhexyl)phthalate
diethylphthalate
dimethylphthalate
2,4-dinitrotoluene
2,6-dinitrotoluene
diphenamid
disulfoton
disulfoton sulfone
disulfoton sulfoxide
endosulfan I
endosulfan II
endosulfan sulfate
0.50 '
• 5.0
0.50
0.50
0.50
0.50
0.50''
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50 '••
ND
0.50 :•
0.50 :
0.50
0.50
ND
0.50
0.50
0.50
o.so :
0.50-
5.0
0.50
0.50
0.50 '
0.50
0.50
0.51
6.5
0.40
0.61
0.55
0.57
0.39
0.71
0.52
0.55
0.54
0.40
0.79
0.41
^0.53
• ND
0.40
0.55
0.48
0.42
ND
0.59
• 0.60
0.60
0.60
0.54
3.99
0.74
0.58
0.55
0.50
0.62
5.7
6.9
7.2
6.2
2.7
6.9
7.0
8.0
6.1
5.8
4.4
6.3
3.5
8.8
0.5
ND
11
9.1
3.7
7.1
ND
9.6
3.2
5.6
8.8
2.5
5.1
3.2
12
18
29
7.2
102
130
80
121
110
113
78
141
104
109
107
80
159
85
106
ND
80
110
96
84
ND
118
120
119
121
107
80
148
116
110
99
124

MDL
C/ig/L)

0.088
1.3
0.086
0.11
0.044
0.12 .'
0.082
0.17
0.095
0.094
0.071
0.075
0.083 .
0.11 •
0.010 '
ND •
0.14
0.15
0.053
0.090-
ND
0.17
0.058
0.099.
0.16
0.041
0.62
0.070
0.20
0.30
0.44
0.13
                                                     525.2-37

-------
TABLE 3.         ACCURACY AND PRECISION DATA  FROM  EIGHT  DETERMINATIONS  OF THE METHOD ANALYTES IN REAGENT WATER USING
                 LIQUID-SOLID C-18 CARTRIDGE EXTRACTION AND THE QUADRUPOLE MASS SPECTROMETER
                 (CONTINUED)


Compound
True
Cone.
Qtg/D
Mean
Observed
Cone'.

-------
TABLE 3.
                 ACCURACY AND PRECISION DATA FROH EIGHT DETERMINATIONS OF THE METHOD  ANALYTES  IM  BPircuT  uirro

                 (CONnNUEDUD C"18 CARTRIDGE EXTRACTI°N AND THE QUADRuPOLE MASS &CTROTCTER             T  "*"*


I Compound

2,2',3,3',4,5',6,6'-octachlorobiphenyl
pebulate
2,2',3',4,6-pentachlorobiphenyl
pentachlorophenol
permethrin, cis
permethrin, trans
phenathrene
prometon'
ppometryn
pronamide
propachlor
propazine
|| pyrene '
simazine
sfmetryn
stirofos : ' . :
II tebuthiuron
terbaci I
terbufos
terbutryn
2,2',4,4'-tetrachlorobiphenyl
toxaphene
- • — r—
tnademefon
2,4,5-trichlorobiphenyl
tricyclazole'
trifluraLin
vernolate

True
Cone.
(MA)

0.50
0.50
0.50
NO
0.25
0.75
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
5.0
5.0
0.50
0.50
0.50
NO
0.50
0.50
5.0
0.50
0.50
1
Mean
Observed
<£g/L)
Relative
Standard
{%)

0.50
0.49
0.30
, . ND
0.30
0.82
0.46
0.30
0.46
0.54
0.49
0.54
0.38
0.55
0.52
0.75
6.8
4.9
0.53
0.47
0.36
ND
0.57
0.38
4.6
0.63
0.51
== '
8.7
5 4
16
ND
Tt -7
2 7
4 3
42
5 6
5 9
•7 r
7.1
5.7
9.1
8.2
5.8
14 •
14
6.1
7 6
4.1
ND
20
6 7
19
5.1
5.5
====i=
Mean Method
Accuracy
<% of True
Cone.)

101
98
61
ND
121
109
92
60

108
98
108

109
105
149
136
97
106
95
71

113

92
127
102

MDL
Gig/D








0 38



0.12
0.066


0 13
2.8





0 33

o
0.096
0.084 1
ND = Not Determined


  Data  from samples extracted at pH 2 - for accurate determination of this analyte  a
  extracted at ambient pH.                                                        '
separate sample must be
                                                    525.2-39

-------
TABLE 4          ACCURACY AND PRECISION DATA FROM EIGHT DETERMINATIONS OF THE METHOD ANALYTES IN REAGENT WATER USING
                 LIQUID-SOLID C-18 DISK EXTRACTION AND THE QUADRUPOLE MASS SPECTROMETER


Compound
True
Cone.

-------
TABLE 4.
ACCURACY AND PRECISION  DATA  FROM EIGHT DETERMINATIONS OF THE METHOD ANALYTES  IN  REAGENT  WATER  USING
LIQUID-SOLID C-18 DISK EXTRACTION AND THE QUADRUPOLE MASS SPECTROMETER (CONTINUED)
• ' ' • ' • - : ' ' ~ 	 ' I

Compound
True
Cone.
Otg/D
Mean
Observed
Cone.

Relative
Standard
Deviation
. {%)
Mean Method
Accuracy
(% of True

chlorneb
chlorobenzilate
2-chlorobiphenyl
chlorpropham
chlorpyrifos
chlorothaConi I
chrysene
cyanazine
cycloate
DCPA
4,4'-DDD
4,4'-DDE
4,4'-DDT
diazinon
dibenz [a, h] anthracene
di-n-butylphthalate
2,3-dichlorobiphenyl
dichlorvos
dieldrin
di - (H-ethylhexyl )adipate
di (2-ethylhexyl )phthalate
diethylphthalate
diniethylphthalate
2,4-dinitrotoluene
2,6-dini trotoluene
diphenamid
disulfoton
disulfoton sulfone
disulfoton sulfoxide
endosulfan I
endosulfan II
1 endosulfan sulfate
0.50
5.0
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
ND
0.50
0.50
0.50
ND
ND
0.50
0.50
0.50
0.50
0.50
5.0
0.50
0.50
0.50
0.50
0.50
0.51
7.9
0.42
0.68
0.61
0.59
0.35
0.68
0.53
0.55
0.67
0.48
0.93
0.56
0.61
ND
0.46
0.54
0.52
ND
ND
0.66
0.57
0.54
0.48
0.60
4.8
0.82
0.68
0.65
0.60
0.67
7.3
8.4
1.9
5.4
6.5
6.5
3.6
15
4.9
4.5
14
4.9
3.2
6.8
' 15
ND
8.1
5.6
7.8
ND
ND
10
8.3
5.7
4.9
3.8
9.4
2.8
8.9
10
21
6:1
100
156
84
134
119
116
71
136
106
110
137
96
187
109
122
ND
93
108
104
ND
ND
132
114
109
96
118
96
164
136
132
122
133
	
MDL


0.11
2.0
0.023
0.11
0.12
0.11-
0.038
0.31
0.077
0.073
0.28
0.070
0.090
0.11
0.28
ND
0.11
0.092
0.12
ND
ND
0.20
0.14
0.093
0.071
0.067
1.3
0.070
0.18
0.20
0.38
0.12
                                                     525.2-41

-------
TABLE 4          ACCURACY AND PRECISIOM DATA  FROM  EIGHT  DETERMINATIONS  OF THE METHOD ANALYTES IN REAGENT WATER USING
                 UQUID-SOUD C-18 DISK EXTRACTION AND THE QUADRUPOLE MASS SPECTROMETER (CONTINUED)


Compound
True
Cone.


0.31
0.24
0.056
0.048
0.090
1.6
1.2
0.11
0.77
0.20
0.28
0.15
0.12
0.14
0.093
0.14
0.12
0.13
0.16
0.14
0.050
0.052
0.033
0.25
0.084
0.062
0.079
0.10
0.030
0.050
0.11
0.089
                                                        525.2-42

-------
TABLE 4.
                ACCURACY  AND PRECISION DATA FROM EIGHT DETERMINATIONS OF  THE  METHOD  ANALYTES IN REAGENT WATER usrur
                LIQUID-SOLID. C-18  DISK EXTRACTION AND  THE QUADRUPOLE MASS  SPECTROMETER  (CONTINUED)
. II

Compound

2,2',3,3',4,5',6,6'-octachlorobiphenvl
pebulate
2,2',3',4,6-pentachlorobiphenyl
pentachlorophenol
permethrin.cis
permethrin, trans
phenathrene
prometon' ; ,
prometryn
pronamide
propachlor
propazine
pyrene
simazine
simetryn
stiro.fos
tebuthiuron
terbacit
terbufos
terbutryn
2,2' ,4,4' -tetrach lorobiphenyt
toxaphene
triademefon
2,4,5-trichlorobiphenyl
tricyclazole
trif luralin
vernolate
True
Cone.
Otg/U

0.50
0.50
0.50
2.0
0.25
0.75
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
5.0
5.0
0.50
0.50
0.50
NO
0.50
0.50
5.0
0.50
0.50
Mean
Observed
Cone.
Oig/D
Relative
Standard
(%)

0.51
0.48
0.35
1.9
0.32
0.89
0.48
0.21
0.46
0.58
0.49
0.59
0.40
0.60
0.41
0.84
9.3
5.0
0.62
0.46
0.40
ND
0.73
0.44
6.8
0.62
0.51
4.2
5.8
4.2
16
3.3
1.9
5.0
66
24
7.1
5.4
5.0
3.2
10
15
3.2
8.6
11
4.2
23
7.4
ND
7.2
5.3
12
2.6

Mean Method
Accuracy
(% of True
Cone.)

102
96
70
95
126
118
95
45
93
113
98
117
79
120
83
168
187
100
123
94

ND
145
89
137
124

	
HDL
(/tg/D

0.064
0.084
0.044
.89
0.031
0.051
0.071
0.44
0.33
0.12
0.079
0.088
0.038
' 0.18
0.19
0.081

1.7
0.077
0.32
0.088
ND
0.16
0.071
2.4
0.048

ND = Not Determined


  Data  from samples extracted at  pH  2 - for  accurate  determination of this analyte   a
  extracted at ambient  pH.
separate sample must be
                                                     525.2-43

-------
TABLE 5          ACCURACY AND PRECISION DATA FROM EIGHT DETERMINATIONS OF THE METHOD ANALYTES  IN REAGENT WATER USING
                 LIQUID-SOLID C-18 CARTRIDGE EXTRACTION AND THE ION TRAP MASS SPECTROMETER


Compound
True
Cone.
C/ig/L)
Mean
Observed
Cone.
((ig/L)
Relative
Standard
Deviation
(%)

Mean Method
Accuracy
(% of True
Cone . )

MDL
C/ig/L >

Surrogates • 	 r
1,3-dimethyl-2-mtrobenzene
perylene-d12
triphenylphosphate
5.0
5.0
5.0
4.9
4.3
4.8
8.4
18
13

98
86
96





Target Analytes 	 T
acenaohthylene
alachlor
aldrin
amctryn
anthracene
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
aroclor 1254*
aroclor 1260
atraton'
atrazine
bcnz ta] anthracene
benzo Cb] f luoranthene
bcnzo Ck] f luoranthene
benzo [g . h , i ] pery I ene
benzo [a] pyrene
bromaci I
butachlor
butyl ate
butylbenzylphthalate"
carboxin
chlordane, (alpha-chlordane)
chlordane, (gamma-chlordane)
chlordane, (trans-nonachlor) 	
0.50
0.50
0.50
0.50
0.50
1.0
ND
ND
ND
ND
1.0 ,
1.0
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
5.0
0.50
0.50
0.50
0.50
• 0.50
0.58
0.4H
0.46
0.42
1.1
ND
'ND
ND
ND
1.1
0.96
0.35
0.55
0.43
0.44
0.34
0.38
0.36
0.45
0.67
0.52
5.7
0.58
0.47
0.50
0.48
8.8
4.0
3.5
3.3
3.8
4.4
ND
ND
ND
ND
17
9.3
11
5.0
7.3
16
22
31
21
9.1
12
5.2
7.7
22
12
10
11

100
115
85
91
84
113
ND
ND
ND
ND
110
96
70
109
85
88
68
76
73
90
133
104
114
117
95
99
96

0.13
0.069
0.045
0.045
0.048
0.15
ND
ND
ND
ND
0.56
0.27
0.12
0.081
0.093
0.21
0.23
0.35
0.23
0.12
0.24
0.082
1.4
0.38
0.17
0.16
0.16
                                                       525.2-44

-------
TABLE 5.        ACCURACY  AND PRECISION DATA FROM EIGHT DETERMINATIONS OF  THE  METHOD  ANAUTES IH REAGENT VATER USIWi
                LIQUID-SOLID C-18  CARTRIDGE  EXTRACTION AND  THE  ION TRAP MASS SPECTROMETER  (CONTINUED)
                                                                                                                           I


Compound
True
Cone;
(/ig/L)

chlorneb
chlorobenzi late
2-ch lorobiphenyl
chlorpropham
chlorpyrifos
chlorothaloni I
chrysene
cyanazine
cycloate
DCPA
4,4'-DDD
4,4'-DDE
4,4'-DDT
diazinon
d ibenz [a, h] anthracene
di-n-butylphthalate"
2, 3 -dich lorobiphenyl
dichlorvos
dieldrin
di (2-ethy Ihexyl )adipate
di(2-ethylhexyl)phthalate"
diethylphthalate
dimethylphthalate
2,4-dinitrotoluene
2,6-dinitrotoluene
diphenamid
disulfoton
disulfoton sulfone
disulfoton sulfoxide
endosulfan I
endosulfan II
endosulfan sulfate
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
5.0
0.50
0.50
0.50
0.50
5.0
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
Mean
Observed
Cone .
C/ig/L)

0.51
0.61
0.47
0.55
0.50
0.62
0.50
0.49
0.52
.. 0.55
0.52
0.41
0.54
0.37
0.37
6.2
0.45
0.53
0.50
0.59
6.5
0.63
0.51
0.45
0.40
0.55
0.62
0.64
0.57
0.60
0.64
0.58
Relative
Standard
Deviation
<%>
Mean Method
Accuracy
(% of True
Cone . )

8.1
9.7 •
4.8
8.1
2.4
5.3
9.2
13
7.6
7.2
3.6
5.8
2.4
2.7
29
4.6
5.8
8.0
10
18
6.6
15
9.5
18
17
6.5
9.8
3.5
8.6
6.1
3.9
5.4
103
123
94
109
99
123
99
97
103
109
103
81
108
75
74
124
90
106
100
117
130
126
102
91
'80
111
124
128
114
121
128
116

HDL

-------
TABLE 5          ACCURACY AND PRECISION DATA  FROM  EIGHT  DETERMINATIONS  OF THE METHOD ANALYTES IN REAGENT WATER USING
                 LIQUID-SOLID C-18 CARTRIDGE EXTRACTION AND THE ION TRAP MASS SPECTROMETER (CONTINUED)


Compound
True
Cone.

-------
TABLE 5.        ACCURACY  AND PRECISION DATA FROM EIGHT DETERMINATIONS OF  THE  METHOD  ANW.YTES IK REAGENT WATER
                LIQUID-SOLID C-18  CARTRIDGE EXTRACTION AND  THE  ION TRAP MASS SPECTROMETER  (CONTINUED}


Compound

2,2' ,3,3' ,4,5' ,6,6'-octachlorobiphenyl
pebulate
2,2' ,3' ,4,6-pentachlorobiphenyl
pentachlorophenol
pennethrin,cis
permethr in, trans
phenanthrene
prometon' . ',
prometryn
pronamide
propachlor
propazine
pyrene
simazine
simetryn
stirofos
tebuthiuron
terbacil
terbufos
terbutryn
2,2',4,4'-tetrachlorobiphenyl
toxaphene
triademefon
2,4,5- trichtorobiphenyl
tricyclazole
trifluralin
vernolate

True
Cone.
C/ig/L)

0.50
0.50
0.50
2.0
0.25
0.75
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50-
0.50
0.50
0.50
0.50
0.50
10
0.50
0.50
0.50
0.50
0.50

Mean
Observed
Cone.

-------
TABLE 6.
ACCURACY AND PRECISION DATA FROM EIGHT DETERMINATIONS OF THE METHOD ANALYTES  IN REAGENT WATER USING
LIQUID-SOLID C-18 DISK EXTRACTION AND THE ION TRAP MASS SPECTROMETER


Compound
True
Cone.
(/tg/L)
Mean
Observed
Cone.


Surrogates
1,3-dfmethyl-2-nitrobenzene
perylene-d12
triphenylphosphate
5.0
5.0
5.0
4.9
4.9
5.9
10
4.5
8.1'
98
98
117




Target Analytes
acenaphthylene
alachlor
aldrin
ametryn
anthracene
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor '248
Aroclor 1254
Aroclor 1260'
atraton"
atrazine
bcnz ta) anthracene
benzotbjfluoranthene
benzolk] f luoranthene
bcnzo[g,h,5]perylene
benzo [a] pyrene
bromacil
butachlor
butyl ate
butylbenzylphthalate0
carboxin
chlordane, (alpha-chlordane)
chlordane, (gamma-chlordane)
chlordane, (trans-nonachlor)
0.50
0.50
0.50
0.50
0.50
0.20
0.20
0.20
0.20
0.20
0.20
0.20
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
5.0
0.50
0.50
' 0.50
0.50
0.51
0.54
0.45
0.41
0.39
0.25
0.26
0.24
0.26
0.24
0.22
0.21
0.10
0.56
0.44
0.50
0.46
0.47
0.44
0.49
0.66
0.50
5.7
0.40
0.50
0.51
0.52
4.5
6.6
6.3
23
15
4.7
6.1
4.7
4.9
4.1
3.7
2.2
46
4.6
7.4
9.1
2.2
i 7.9
; 12
4.4
5.1
! 5.4
: 7.7
38.1
4.3
7.2
6.2
102
108
90
82
79
123
130
121
129
118
110
108
21
111
88
100
91
95
89
99
132
100
114
79
101
102
104
0.068
0.11
0.085
0.29
0.18
0.040
0.054
0.042
0.043
0.038
0.028
0.018
0.14
0.076
0.098
0.14
0.031
0.11
0.16
0.066
0.10
0.082
1.4
0.45
0.065
0.11
0.097
                                                     525.2-48

-------
TABLE 6.         ACCURACY  AND PRECISION DATA FROM EIGHT DETERMINATIONS OF  THE  METHOD•ANALYTES IN REAGENT WATER'USING
                 LIQUID-SOLID C-18  DISK EXTRACTION AMD  THE  ION  TRAP MASS  SPECTROMETER  (CONTINUED)


Compound
True
Cone.
to/L)
Mean
Observed
Cone.
<^g/L) ...
Relative
Standard
Deviation
(%)
Mean Method
Accuracy
(% of True
Cone . )

MDL
Oig/L>

chlorneb
chlorobenzilate
2-chlorobiphenyl
chlorpropham
chlorpyrifos
chlorothaloni I
chrysene
cyanazine
cycloate
DCPA
4,4'-DDD
4, 4' -DDE
4,4'-DDT
diazinon
dibenz [a, h] anthracene
di-n-butylphthalate"
2,3-dichlorobiphenyl
dichtorvos
dieldrin
di (2-ethy Ihexyl )adipateB
di(2-ehtylhexyl)phthalate"
diethylphthalate
dimethylphthalate
2,4-dinitrotoluene
2,6-dinitrotoluene
diphenamid
disulfoton
disulfoton sulfone
disulfoton sulfoxide
endosutfan I
endosulfan II
endosulfan sulfate
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
5.0
0.50
0.50
0.50
5.0
5.0
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.54
0.59
0.50
0.55
0.54
0.59
0.48
0.52
0.51
0.53
0.63
0.48
0.58
0.50
0.47
5.7
0.50
0.50 .
0.53
5.4
5.7
0.68
0.51
0.30
0.28
0.56
0.70
0.64
. 0.60
0.61
0.66
0.57
6.3
9.7
4.7
4.7
11
4.4
6.1
8.3
4.1
3.2
16
3.7
7.2
4.5
9.9
3.3
2.6
8.7
7.0
7.5
2.6
5.0
5.0
8.1
6.4
6.4
5.3
5.9
3.8
4.9
6.1
9.0
108
117
100
111
109
119
96
105
102
105
127
96
117
101
94
115
100
99
106
107
114
137
102
59
56
112
139
128
119
122
131
115
0.10
0.17
0.070
0.079
0.18
0.079
0.088
0.13
0.063
0.051
0.31
0.054
0.13
0.068
0.14
0.59
0.039
0.13
0.11
1.3
0.46
0.10
0.077
0.072
0.054
0.11
0.11
0.11
0.068
0.089
0.12
0.16
                                                      525.2-49

-------
TABLE 6.
ACCURACY AND PRECISION  DATA  FROM EIGHT DETERMINATIONS OF THE METHOD ANALYTES  IN  REAGENT  WATER  USING
LIQUID-SOtID C-18 DISK EXTRACTION AND THE ION TRAP MASS SPECTROMETER (CONTINUED)
• -

Compound

endrfn
cndrin aldehyde
EPTC
cthoprop
etridiazole
fenacniphos
fcnarlmol
fluorane
fturidone
HCH, alpha
HCH, beta
HCH, delta
HCH, garama (lindane)
hcptachlor
hcptachlor epoxide
2,2',3,3',4,4',6-heptachlorobipheny
I
hcxach lorofaenzene
2,2',4,4',5,6'-hexachlorobiphenyl
hexach lorocyc I opentadi ene
hexazinone
{ndenoC1,2,3-cd]pyrene
isophorone
mcthoxychlor
methyl paraoxon
mctolachlor
metribuzin
mrvinphos
HuK 264 • isomer a
HGK 264 - isomer b
molinate
napropamide
norflurazon

True
Cone.
(M/L)

0.50
0.50
0.50
0.50
0.50
0.50
0.50
0,50 -•
5.0
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.33
, 0.16
0.50
0.50
' 0.50

Mean
Observed
Cone.
(jig/D

0.68
0.57
0.48
0.61
0.54
0.67
0.59
0.53
5.2
Q.55
0.54
0.53
0.50
0.49
0.50
0.46
0.49
0.50
0.37
0.75
0.48
0.51
0.52
0.75
0.57
0.53
0.56
0.38
0.18
0.53
0.58
0.71

Relative
' Standard
: Deviation
C%)

7.9
2.8
5.2
.:•• 7.5
4.2
10
5.8
•i • 3.4
2.3
5.0
4.1
3.6
3.2
4.0
. 3.2
I
• 7.3
3.4
1 5.3
9.3
4.2
7.3
i 4.3
6.7
.4.5
3.2
5.7
6.2
6.7
5.3
3.8
7.9
4.3

Mean Method
Accuracy
(% of True
Cone.)

137
114
97
122
108
133
118
• 106
104
110
109
106
100
98
100
• • 92
97
99
73
150
96
102
104
151
114
107
112
113
110
105
116
142

MDL
Oig/L)

0.16
0.048
0.076
0.14
0.067
0.20
0.10
0,054
0.16
0.083
0.068
0.058
0.047
0.059
0.048
0.10
0.049
0.079
0.10
0.094
0.10
0.066
0.10
0.10
0.054
0.090
0.10
0.076
0.029
0.060
0.14
0.091
                                                       525.2-50

-------
TABLE 6.
ACCURACY: AND PRECISION DATA  FROM  EIGHT DETERMINATIONS OF. THE METHOD ANALVTES  IN  REAGENT  WATER USING
LIQUID-SOLID C-18 DISK EXTRACTION AND THE ION TRAP MASS SPECTROMETER (CONTINUE0)


Compound

2,2',3,3',4,5',6,6'-
octachlorobiphenyl
pebulate
2,2' ,3' ,4,6-pentachlorobiphenyl
pentachlorophenol
permethrin,cis
permethrin, trans
phenanthrene
prometon"
prometryn
pronamide '
propachlor
propazine
pyrene
simazine
simetryn
stirofos
tebuthiuron
terbaci I
terbufos
terbutryn
2,2',4,4'-tetrachlorobiphenyl
toxaphene'
triademefori
, 2,4,5-trichlorobiphenyt
tricyclazole
trifluralin
vernolate
True
Cone.
Oig/U
Mean
Observed
Cone.
(jitg/L)
Relative
Standard
(%)
Mean Method
Accuracy
(% of True

0.50
0.50
0.50
2.0
0.25
0.75
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50 '
0.50
. 0.50
0.50
0.50
.10
0.50
0.50-
0.50
0.50
0.50
0.47 '
0.56
0.49
2.2
0.37
0.84
0.49
0.16
0.46
0.56
0.58
0.53 ,
0.52
0.54
, 0.36 ,
0.72
0.67
0.64
0.57
0.46
0.46
12
0.71
0.48
0.65
0.59
0.50
5.3
7.1
4.0
15
3.1
1.6
6.3
63
23
3.9
5.7
4.7
5.2
2.8
20
3.7
7.9
12
6.8
24
7.4
2.7
7.3
4.5
14
7.8 .
3.2
94
112
97
111
149
112
97
32
91
111
115
106
104
107
, 71 :
144
133
129
113
93
91
122
142
97
130
117
99
.)
MDL
(^9/L)

0.076
0.11
0.059
1.0
0.035
0.039
0.092
0.30
0.32
0.064
0.098
0.074
0.080
0.045
0.22
0.080
0.16
0.23
0.11
Oi34
0.10
1.0
0.16
0.066
0.27
0.14

   Six replicates


   Seven replicates in fortified tap water.                                               '            *


   Seven replicates


   Data from samples  extracted  at pH 2  -  for  accurate determination of  this  analyte.  a separate sample  must  be
   extracted at ambient pH.                                                                                     •
                                                     525.2-51

-------
S
i
co :
u '



































































0
*J
§
c.
u

CO

C/)
(0

cu
^

•O
ro
3
O











t..
O
+••
t_
O
C/)
en
ra
X
Q.
CD
L.
I—
g






|



U









.jjf
(A
'•"












O)
TD
t_
L.
to
CJ










W
O






OJ
Dl
T3
L.
+•*
L.
S











•O 0)
0 3
.C >. t- >*
4-» O h- •
o ro o
Z t- H- C
3 O O
CO O
CO O S£
0) < v^
3C

c
4) "D O
.i ro '-M ^
4-" TJ 10 S«

0) 4J 01
at tn a

~O 0)
0 3
^ >• I- *-*
4-> O 1— •
Ol (0 U
31 3 'o 0
CO CJ
ra o s«
01 

0) T3 o
•>ro«^
4J "D ro te
^
— ro >
Ol 4-- OJ
o: co Q
"8 3
J= >. t- •<-»
4J O h- •
O) ro o
3C t— **• C
3 O O
CO U
ro o 3-S
0) *C ^
z
c
SI'S .2
•i— m 4-» *+
*j "b re oa
CD C •«- ^-^
SSS
"U cu
0 3
^ >• ^ <*
•M O If— •
0) CO U
Z *- H- C
3 O O

(U < vx
OJ C
>T3 0
m ? .2 S
or *j 01
CO O





































































OJ








CO
CO






8





•o




g






tn
-*



8;


N-
N-



w



. 'g

S
H-



O
tn








in
\A






-O
(M





in




o







o



o


o
(M







c_

H-



V—








in
r»






(^
O


,

.
•0
io




S






r°
PJ
. i

i

in
o
L
|
in
(M







2
'u
1



N








ro
CO






o





in




S






j^
c>



>o
o



-*



(U


c
N
(D
X
01
.c



ro








^
«-






»





ro




Ox










CO








CO
o>






o
o





r-
ro




CO
CM







o



ro


00
f—



i
4-»


TJ

C.
4->



N-
ro









PO






OJ
o





S




o
o






in
o:



CO



•o


01
"o
N


°^
O
L.
                            525.2-52

-------
TABLE 8.
              ACCURACY AND PRECISION DATA FROM SEVEN DETERMINATIONS OF THE METHOD ANALYTES [H
              TAP WATER USING LIQUID-SOLID C-18 CARTRIDGE EXTRACTION AND THE ION TRAP MASS
              SPECTROMETER                                                          ;

Compound

acenaphthylene
alachlor
aldrin
ametryn
anthracene
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroctor 1242
Aroclor 1248
APbclbr 1254
Aroclor 1260 . , , -'....
atraton'
atrazine
benz [a] anth racene
benzo[b]f luoranthene
t-:nzo[k]f luoranthene
benzo[g,h,i]perylene
benzo Ca] pyrene
bromacil
butachlor
butyl ate . . :
butylbenzylphthalate
carboxin
chlordane, (alpha-chlordane)
chlordane, (gamma-chlordane)
chlordane, (trans-nonachlbr)
chlorneb
chlorobenzi late
2Jchlorobiphenyl
chlorpropham
chlorpyrifos
chlorthalonil
chrysene

True Cone*

5.0
5.0
5.0
5.0
5.0
ND
ND
ND
ND
ND
. ' ND
ND
• . 5.0 •
5.0
' 5.0.
'S.O
5.0
'5.0
5.0
5.0
5.0
5.0
5.0
5.0
'5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0

Mean

5.2
5.5 „
4.4
4.2
4.3
ND
ND
ND
ND
ND
ND
ND
2.2 ,
5.6
4.9
5.7
5.7
5.6
6.1
3.5
5.4
5.1
7.2
1.0
5.2
5.1
5.6
5.2
5.7 ;
5.8
6.3
5.3
5.4
5.5

% RSD
.'.-
5.3
6.9
14
3.4
5.2'
ND
ND
ND
ND
ND
ND
• ND
28
6.2
8.8
7.5
2.9
7.1
4.6 .
5.1
'7.5
4.5 •;
8.3
23
8.9
8.0
7.4
3.0 '
4.4
5.4
4.9
7.2
9.9
3.9

% REC

104
110
88
83
87
ND
. ND
ND
ND
ND
ND
ND
43
111
97
114
113
113
121
69
109
102
144
20
104
102
111
105
114
115
127
107
108
110
                                        525.2-53

-------
TABLE 8.      ACCURACY AND PRECISION DATA FROM SEVEN DETERMINATIONS OF THE METHOD ANALYTES IN TAP WATER
              USING LIQUID-SOLID C-18 CARTRIDGE EXTRACTION AND THE ION TRAP MASS SPECTROMETER (CONTINUED)


Compound

cyanazine
cycloate
DCPA
4,4'-DDD
4,4'-DDE
4,4'-DDT
diazinon
dibenz [a, h] anthracene
di-n-butylphthalate
2,3-dichlorobiphenyl
dichlorvos
dietdrin
di (2-ethylhexyl )adipate
di(2-ethylhexyl)phthalate
diethytphthalate
dfmethylphthalate
2,4-dinitrotoluene
2,6-dinitrotoluene
diphenamid
disutfoton
disulfoton sulfone
disulfoton su If oxide
endosulfan I
endosulfan II
endosulfan sulfate
endrin
endrin aldehyde
EPTC
ethoprop
etrfdiazole
fenamfphos
fenarimol
f luorene
fluridone
HCH, alpha
True Cone.

5.0
5.0 •
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0;
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
s;o
s.o ;
Mean

6.1
5.6
5.4
5.3
5.2
5.6
4.9
5.9
6.2
5.3
2.8
5.3
6.7
6.5
6.4
5.8
4.2
4.1
5.2
2.5
5.5
9.4
5.5
5.3
5.3
6.1
5.1
5.1
6.3
5.8
5.9
7.1
5.7
6,2
5.9
% RSD

13
1.5
5.0
6.5
6.6
9.6
8.7
7.5
4.6
7.4
7.3
1 7.2
10
6.6
7.4
7.1
8.7
8.5
7.7
33
7.4
11
11
9.6
7.8
3.9
9.1
2.1
4.2
7.5
22
3.3
5.2
9.0
2.6
% REC

122
112
107 .
105
104
111
98
118
124
106
56
105
134
130
. 127
116
84
82
104
50
110
188
109
106
106
121
102
102
125
117
119
141
114
125
118
                                                  525.2-54

-------
TABLE 8.      ACCURACY AND PRECISION DATA FROM SEVEN DETERMINATIONS OF THE METHOD ANALYTES IN TAP UATEK
              USING LIQUID-SOLID C-18 CARTRIDGE EXTRACTiON AND THE ION TRAP MA'SS SPECTROMETER (CONTINUED)

Compound

HCH, beta
HCH, delta
HCH, gamma (Lindane)
heptachlor
heptachlor epoxide
2,2' ,3,3' ,4,4' ,6-heptachlorobiphenyl
hexach I orobenzene
2,2',4,4',5,6'-hexachlorobiphenyl
hexachlorocyclopentadiene •
hexazinone
i ndeno [ 1 , 2 , 3 - cd] py rene
isophorone
methoxychlor
methyl paraoxon
;
metolachlor
metribuzin •
mevinphos
MGK 264 - isomer a
MGK 264 - isomer b
molinate
napropamide
norf turazon '.
2,2',3,3',4,5',6,6'-octactorobiphenyl
pebulate
2,2' ,3',4,6-pentachlorobiphenyl
pentach I oropheno I
permethrin, cis
permethrin, trans
phenanthrene
prometona*.
prometryn
pronamide
propachlor
propazine
pyrene

True Cone.

5.0
5.0
5.0
5.0
•• 5.0
5.0
5.0
5.0
5.0 .
5.0
5:0
5.0
- 5.0
5.0
5.0
5.0
5.0
3.3
1.7
5.0
5.0
-5.0
5;0
• 5.0
5.0
20.
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0 .•
5.0

Mean

5.3
5.3
5.3
4.7
5.2
5.1
4.6
5.6
6.0
6.9
6.8
4.9
5.6
5.6
5.6
2.1
3.3
3.6
1.8
5.5
5.3
6.7
4.9
5 .3
5.3
33
3.3
8.5
5.5
2.0
4.5
5.7
6.2
5.6
5.2

% RSD

8.4
5.2
6.9
8.7
. 7.7
6.9
7.4
8.1
4.8
6.3
7.7
12
4.9
11
7.7
5.8
1.6
6.2
7. -6
1.5
8.9
7.2
6.9
3.1
8.1
4.9
3.5
2.2
4.0
25
4.3
5.3
4.0
4.9
6.7

% REC

106
106
107
93
105
. 103
93
112
120
138
135
99
112
111
111
42
67
107
110
, 110
106
135
97
106
107
162
130
113
109
40
89
115
124
' 113
104
                                                  525.2-55

-------
TABLE 8.
ACCURACY AND PRECISION DATA FROM SEVEN DETERMINATIONS OF THE METHOD ANALYTES IN  TAP WATER
USING LIQUID-SOLID C-18 CARTRIDGE EXTRACTION AND THE ION TRAP MASS SPECTROMETER  (CONTINUED)
•
Compound

simazine
simetryn
stirofos
tebuthiuron
terbaci I
terbufos
terbutryn
2,2',4,4'-tetrachlorobiphenyl
toxaphene
triademefon
2,4,5-trichlorobiphenyl
tricyclazole
trif lutelin
vernotate
True Cone.

5.0
5.0
5.'o
5:0
5.0
5 JO
5.0
5.0
i,ND
5 JO
5 Jo
5.0
5JO
5JO
Mean

6.0
3.9
6.1
6.5
4.0
4.5
4.3
5.3
ND
6.0
5.2
4.8
5.9
5.4
% RSD

9.0
7.0
12
9.7
5.5
8.4
6.5
4.3
ND
12
5.1
5.2
7.8
3.3
% REC

120
78
121
130
79
90
86
106
ND
121
103
96
119
108
     Data from samples extracted at pH 2 - for accurate determination of this analyte,  a separate sample
     must be extracted at ambient pH.
                                                  525.2^56

-------
                     2 Liter
                   separator/
                     funnel
                                   K3
      125ml
      solvent
     reservoir

     ground glass T 14/35

     IS£ cartridge

J y rubber stopper


     No. 18-2O tuer-lok
        syringe needle
                    1 fiter
                 vacuum flask
                                                      125 ml
                                                      solvent
                                                      reservoir

                                                      ground glass
                                                        I 14/35
                                                    ISE cartridge
                                             10Oml
                                           separator/
                                             funnel
                                                       drying
                                                      column
                                                      (Na,SOJ
                                                   0.6 cm x 4O cm
                                                     1O ml
                                                 f graduated
                                                 I     vial
A. Extraction apparatus
                           B. Etutlon apparatus
     FIGURE 1.   CARTRIDGE EXTRACTION APPARATUS
                            525.2-57

-------
525.2-58

-------
METHOD 531.1.  MEASUREMENT OF N-METHYLCARBAMOYLOXIMES AND N-METHYLCARBAMATES
               IN WATER BY DIRECT AQUEOUS  INJECTION HPLC WITH POST COLUMN
               DERIVATIZATION
                                 Revision 3.1
                          Edited by J.W,  Munch (1995)
D.L. Foerst - Method 531, Revision 1.0 (1985)

T. Engel (Battelle Columbus Laboratories) - National Pesticide Survey
         Method 5, Revision 2.0 (1987)

R.L.. Graves - Method 531.1, Revision 3.0 (1989)
                     NATIONAL  EXPOSURE RESEARCH  LABORATORY
                      OFFICE OF RESEARCH AND DEVELOPMENT
                     U.S. ENVIRONMENTAL PROTECTION AGENCY
                            CINCINNATI,  OHIO   45268


                                    531.r-i

-------
                                  METHOD  531.1

                     MEASUREMENT  OF  N-METHYLCARBAMOYLOXIMES
       AND N-METHYLCARBAMATES  IN WATER BY DIRECT AQUEOUS  INJECTION HPLC
                        WITH POST COLUMN DERIVATIZATION


1.   SCOPE AND APPLICATION

     1.1  This is  a  high  performance  liquid chromatographic  (HPLC) method
          applicable to the determinations of certain N-methylcarbamoyloximes
          and N-methylcarbamates in ground water and finished drinking water.
          The following compounds can be determined using this method:

                                         Chemical Abstract Services
              Analvte                           Registry Number	

          Aldicarb                                116-06-3
          Aldicarb sulfone                   \    1646-88-4
          Aldicarb sulfoxide                 .    1646-87-3
          Baygon                                  114-26-1
          Carbaryl                                 63-25-2
          Carbofuran                            1563-66-2
          3-Hydroxycarbofuran                   16655-82-6
          Methiocarb                            2032-65-7
          Methomyl                              16752-77-5
          Oxamyl                            .  .  23135-22-0

     1.2  This method has been validated in a single laboratory and estimated
          detection  limits (EDLs) and method detection limits (MDLS) have been
          determined for  the analytes above.  Observed detection limits may
          vary between ground waters, depending upon the nature of
          interferences in the sample matrix and the specific instrumentation
          used.

     1.3  This method is  restricted to use by or under the supervision of
          analysts experienced in the use of liquid chromatography and in the
          interpretation  of liquid  chromatograms.  Each analyst must demon-
          strate the ability to  generate acceptable results with this method
          using the  procedure described in Sect. 9.3.

     1.4  When this  method is used  to analyze unfamiliar samples for any or
          all of the analytes above, analyte identifications should be
          confirmed  by at least  one additional qualitative technique (1).

2.   SUMMARY OF METHOD                       ;

     2.1  The water  sample is filtered and a 400-/zL aliquot is injected into a
          reverse phase HPLC column.  Separation of the analytes is achieved
          using gradient elution  chromatography.  After elution from the HPLC
          column, the analytes are  hydrolyzed with 0.05 N sodium hydroxide
          (NaOH)  at  95°C.   The methyl amine formed during hydrolysis is

                                    531.1-2

-------
          reacted with o-phthalaldehyde (OPA) and 2-mercaptoethanol to form a
          highly fluorescent derivative which is detected by a fluorescence
          detector (2).  Analytes are quantitated using procedural standard
          calibration  (Sect. 3.14)

3.   DEFINITIONS

     3.1  INTERNAL STANDARD — A pure analyte(s) added to a solution in known
          amount(s) and used to measure the relative responses of other method
          analytes and surrogates that are components of the same solution.
          The internal standard must be an analyte that is not a sample-
          component.

     3.2  SURROGATE ANALYTE — A pure analyte(s), which is extremely unlikely
          to be found in any sample, and which is added to a sample aliquot in
          known amount(s) before extraction and is measured with the same
          procedures  used to measure other sample components.   The purpose of
          a surrogate analyte is to monitor method performance with each
          sample.

     3.3  LABORATORY  DUPLICATES (LD1 and LD2) — Two sample aliquots taken in
          the analytical  laboratory and analyzed separately with identical
          procedures.  Analyses of LD1 and LD2 give a measure  of the precision
          associated  with laboratory procedures,  but not with  sample
          collection, preservation,  or storage procedures.

     3.4  FIELD DUPLICATES (FD1 and FD2) — Two separate samples collected at
          the same time and place under identical  circumstances and treated
          exactly  the same throughout field and laboratory procedures.
          Analyses of FD1 and FD2 give a measure of the precision associated
          with sample collection,  preservation and storage,  as well  as  with
          laboratory  procedures.

     3.5e  LABORATORY  REAGENT BLANK (LRB) — An aliquot of reagent water that
          is treated  exactly as a sample including exposure to all  glassware,
          equipment,  solvents,  reagents, internal  standards, and surrogates
          that are used with other samples.   The  LRB is used to determine if
          method analytes or other interferences  are present in the laboratory
          environment,  the reagents,  or the apparatus.

     3.6  FIELD REAGENT BLANK (FRB)  — Reagent water placed  in a sample
          container in the laboratory and  treated  as a sample  in all  respects,
          including exposure to sampling site conditions,  storage,
          preservation and all  analytical  procedures.   The  purpose  of the FRB
          is to, determine if method  analytes  or  other  interferences  are
          present  in  the  field  environment.

     3.7  LABORATORY  PERFORMANCE  CHECK SOLUTION  (LPC)  —  A  solution  of  method
          analytes, surrogate compounds,  and  internal  standards used  to
          evaluate  the performance of the  instrument system with  respect  to a
          defined  set of  method criteria.
                                   531.1-3

-------
     3.8  LABORATORY  FORTIFIED  BLANK  (LFB) — An aliquot of reagent water to
          which  known  quantities  of the method analytes are added in the
          laboratory." The  LFB  is  analyzed exactly like a sample, and its
          purpose  is  to determine  whether the methodology is in control, and
          whether  the  laboratory  is capable of making accurate and precise
          measurements at the required method detection limit.

     3.9  LABORATORY  FORTIFIED  SAMPLE MATRIX (LFM) — An aliquot of an
          environmental sample  to  which known quantities of the method
          analytes  are added in the laboratory.  The LFM is analyzed exactly
          like a sample, and its  purpose is to determine whether the sample
          matrix contributes bias  to the analytical results.  The background
          concentrations of the analytes in the sample matrix must be
          determined  in a separate aliquot and the measured values in the LFM
          corrected for background concentrations.

     3.10 STOCK STANDARD SOLUTION  — A concentrated solution containing a
          single certified  standard that is a method analyte, or a
          concentrated solution of a single analyte prepared in the laboratory
          with an  assayed reference compound.  Stock standard solutions are
          used to  prepare primary  dilution^ standards.

     3.11 PRIMARY  DILUTION  STANDARD SOLUTION — A solution of several analytes
          prepared  in the laboratory from stock standard solutions and diluted
          as needed to prepare calibration1 solutions and other needed analyte
          solutions.                   •   •-   .     '

     3.12 CALIBRATION STANDARD  (CAL) — A solution prepared from the primary
          dilution  standard solution and s[tock standard solutions of the
          internal  standards and  surrogate analytes.  The CAL solutions are
          used to  calibrate the instrument response with respect to analyte
          concentration.

     3.13 QUALITY  CONTROL SAMPLE  (QCS) — A sample matrix containing,method
          analytes  or a solution of method analytes in a water miscible
          solvent which is  used to fortify reagent water or environmental
          samples.  The QCS is obtained from a source external to the
          laboratory, and is used  to check laboratory performance with
          externally prepared test materials.

     3.14 PROCEDURAL STANDARD CALIBRATION —  A calibration method where
          aqueous calibration standards are prepared and processed (e.g.
          purged, extracted, and/or derivajtized) in exactly the same manner as
          a sample.  All steps in  the process from addition of sampling
          preservatives through instrumental  analyses are included in the
          calibration.  Using procedural standard calibration compensates for
          any inefficiencies in the processing procedure.

4.   INTERFERENCES

     4.1  Method interferences may be caused by contaminants in solvents,
          reagents, glassware and  other sample processing apparatus that lead

                                    531.1-4

-------
           to discrete artifacts or elevated baselines in liquid chromatograms.
           Specific sources of contamination have not been identified.  All
           reagents and apparatus must be routinely demonstrated to be free
           from interferences under the conditions of the analysis by running
           laboratory reagent blanks as described in Sect. 9.2.
           4.1.1
           4.1.2
        Glassware must be scrupulously cleaned.(3)   Clean all
        glassware as soon as possible after use  by  thoroughly rinsing
        with the last solvent used in it.   Follow by washing with hot
        water and detergent and thorough rinsing with tap and reagent
        water.   Drain dry,  and heat in an  oven or muffle furnace at
        40CTC for 1 hour.  Do not heat volumetric glassware.
        Thermally stable materials might not be  eliminated by this
        treatment.   Thorough rinsing with  acetone may be substituted
        for the heating.   After drying.and cooling,  seal  and store
        glassware in a clean environment to prevent  any accumulation
        of dust or  other contaminants.   Store inverted or capped with
        aluminum foi1.

        The use of  high purity reagents  and solvents helps to
        minimize interference problems.   Purification of solvents by
        distillation in all-glass systems  may be  required.   WARNING:
        When a  solvent  is purified,  stabilizers  added by  the
        manufacturer are  removed,  thus potentially making the  solvent
        hazardous.   Also,  when a solvent is purified,  preservatives
        added by the manufacturer are removed, thus  potentially
        reducing the shelf-life.
       .2   Interfering contamination may occur when  a  sample containing
           concentrations of  analytes  is analyzed  immediately following
           sample containing  relatively high ,concentrations of analytes
           preventive technique  is between-sample  rinsing of the sample
           and filter holder  with two  portions of  reagent water.  After
           analysis ,of a sample  containing high concentrations of analytes,
           or more laboratory reagent  blanks should  be analyzed.
                                                              low
                                                              a
                                                               A
                                                              syringe
                                                                 one
     4.3  Matrix interference may be caused by contaminants that are present
          in the.sample.  The extent of matrix interference will vary consid-
          erably from source to source, depending upon the water sampled.
          Analyte identifications must be confirmed.  Positive identification
          may be made by the use of an alternative detector which operates on
        .. a chemical/physical principle different from that originally used;
          e.g., mass spectrometry, or the use of a second chromatography
          column.  A suggested alternative column is described in Sect. 6.6.3.
5.   SAFETY
     5.1
The toxicity or carcinogenicity of each reagent used in this method
has not been precisely defined; however, each chemical  compound must
be treated as a potential health hazard.  Accordingly,  exposure to
these chemicals must be reduced to the lowest possible level.  The
laboratory is responsible for maintaining a current awareness file
                                    531.1-5

-------
     of OSHA regulations regarding the safe handling of the chemicals
     specified in this method.  A reference file of material  safety data
     sheets should also be made available to all personnel  involved in
     the chemical analysis.  Additional references to laboratory safety
     are available and have been identified (4-6) for the information of
     the analyst.

5.2  WARNING:  When a solvent is purified, stabilizers added by the
     manufacturer are removed, thus potentially making the solvent
     hazardous.                       ;

EQUIPMENT AND SUPPLIES  (All specifications are suggested.   Catalog
numbers are included for illustration only.)

6.1  SAMPLING EQUIPMENT               '.

     6.1.1  Grab sample bottle — 60-mt screw cap vials (Pierce No. 13075
            or equivalent) and caps equipped with a PTFE-faced silicone
            septa (Pierce No. 12722 or equivalent).  Prior to use, wash
            vials and septa as described in Sect. 4.1.1.

6.2  BALANCE — Analytical, capable of accurately weighing to the nearest
     0.0001 g.

6.3  FILTRATION APPARATUS

     6.3.1  Macrofiltration — To filter derivatization solutions and
            mobile phases used in HPLC.  Recommend using 47 mm filters
             (Mi Hi pore Type HA, 0.45 pm for water and Mi Hi pore Type FH,
            0.5 /im for organics or equivalent).

     6.3.2  Microfiltration — To filter samples prior to HPLC analysis.
            Use  13 mm filter holder (Millipore  stainless steel XX300/200
            or equivalent), and 13 mm diameter  0.2 p.m polyester filters
             (Nuclepore 180406 or equivalent).

6.4  SYRINGES AND SYRINGE VALVES

     6.4.1  Hypodermic syringe — 10-mL glass,  with Luer-Lok tip.

     6.4.2  Syringe valve — 3-way (Hamilton HV3-3 or equivalent).

     6.4.3  Syringe needle — 7 to 10-cm long,  17-gauge, blunt tip.
     6.4.4  Micro syringes — various

6.5  MISCELLANEOUS
sizes.
     6.5.1  Solution storage bottles — Amber glass, 10- to 15-mL
            capacity with TFE-fluorocarbon-lined screw cap.

     6.5.2  Helium, for degassing  solutions and solvents.

                               531.1-6

-------
 6.6   HIGH  PERFORMANCE  LIQUID  CHROMATOGRAPH  (HPLC)

      6.6.1   HPLC  system  capable  of injecting  200-  to  400-juL  aliquots,  and
             performing binary linear  gradients  at .a constant flow  rate.
             A  data  system  is  recommended  for  measuring  peak  areas.  Table
             1  lists  retention times observed  for method analytes using
             the columns  and analytical  conditions  described  below.

      6.6.2   Column  1  (Primary column) —  150  mm long  x  3.9 mm I.D.
             stainless  steel packed with 4 /mi  NovaPak  CIS.  Mobil Phase  is
             established  at 10:90 methanol:water, hold 2 min.,  then
             program  as a linear  gradient  to 80:20  methanol:water in 25
             min.  Alternative columns may be  used  in  accordance with the
             provisions described in Sect. 9.4.

      6.6.3   Column 2  (Alternative  column)* — 250  mm  long x  4.6 mm I.D.
             stainless  steel packed with 5 /im  Beckman  Ultrasphere ODS.
             Mobile phase is established at 1.0  mL/min as a linear
             gradient from  15:85 methanol:water  to  100 % methanol in 32
             min.  Data presented in this method were  obtained  using this
             column.  * Newer  manufactured columns  have  not been able to
             resolve  aldicarb  sulfone  from oxamyl.

      6.6.4   Column 3 (Alternative  column) —  250 mm long x 4.6 mm  I.D.
             stainless  steel packed with 5 jLim  Supelco  LC-1.   Mobile phase
             is established at  1.0  mL/min as a linear  gradient  from 15:85
             methanol:water to  100  % methanol  in 32 min.

      6.6.5   Post column  reactor — Capable of mixing  reagents  into the
             mobile phase.  Reactor should be constructed using PTFE
             tubing and equipped with pumps to deliver 0.1 to  1.0 mL/min
             of each reagent;  mixing tees;  and two  1.0-mL delay coils, one
             thermostated at 95°C (ABI URS 051 and URA 100 or equivalent).

      6.6.6   Fluorescence detector  — Capable of excitation at 330 nm
             (nominal)  and detection of emission energies greater than 418
             nm.  A Schoffel Model  970 fluorescence detector was used to
             generate the validation data presented in this method.

REAGENTS AND STANDARDS — WARNING:   When a solvent is purified,
stabilizers  added by the manufacturer are removed,  thus potentially
making the solvent hazardous.   Also, when a solvent is  purified,
preservatives added by the manufacturer are removed,   thus potentially
reducing the shelf-life.

7.1  REAGENT WATER —  Reagent water is defined  as water that is
     reasonably free of contamination that would prevent the
     determination of  any analyte  of interest.  Reagent water used  to
     generate the validation data  in this  method was  distilled water
     obtained from the Magnetic Springs Water Co.,  1801 Lone Eagle  St.,
     Columbus,  Ohio 43228.
                               531.1-7

-------
 7.2   METHANOL  —  Distilled-in-glass  quality  or  equivalent.

 7.3   HPLC  MOBILE  PHASE

      7.3.1   Water —  HPLC  grade  (available from Burdick  and Jackson).

      7.3.2   Methanol  —  HPLC  grade.   Filter  and degas with helium  before
             use.

 7.4   POST  COLUMN  DERIVATIZATION  SOLUTIONS

      7.4.1   Sodium hydroxide,  0.05 N  —  Dissolve 2.0 g of sodium
             hydroxide (NaOH)  in  reagent.water.   Dilute to 1.0  L with
             reagent water.  Filter and degas with  helium just  before use.

      7.4.2   2-Mercaptoethanol  (1+1) — Mix 10.0 mL of 2-mercaptoethanol
             and 10.0  mL  of  acetonitrile.  Cap.   Store in hood  (CAUTION —
             stench).

      7.4.3   Sodium borate  (0.05  N) — Dissolve  19.1 g of sodium borate
             (Na2B407 ' 10H20) in reagent water.   Dilute to 1.0 L with
             reagent water.  The  sodium borate will completely  dissolve at
             room  temperature  if  prepared a day  before use.

      7.4.4   OPA reaction solution —  Dissolve 100  + 10 mg of o-phthal-
             aldehyde  (mp 55-58°C) in  10  mL of methanol.  Add to 1.0 L of
             0.05  N sodium borate.  Mix,  filter,  and degas with helium.
             Add 100 juL of 2-mercaptoethanol  (1+1)  and mix.  Make up fresh
             solution  daily.

7.5  MONOCHLOROACETIC'ACID  BUFFER (pH3)  — Prepare by mixing 156 mL of
     2.5 M monochloroacetic acid and  100 mL 2.5  M  potassium acetate.

7.6  4-BROMO-3,5-DIMETHYLPHENYL  N-METHYLCARBAMATE  (BDMC) — 98% purity,
     for use as internal standard (available from Aldrich Chemical Co.).

7.7  STOCK STANDARD SOLUTIONS  (1.00 /tg/jil) — Stock standard solutions
     may be  purchased as certified solutions or  prepared from pure
     standard materials using the following procedure:

     7.7.1   Prepare stock standard solutions by  accurately weighing
             approximately 0.0100 g of pure material.   Dissolve the
            material  in methanol  and dilute to volume in a 10-mL
            volumetric flask.   Larger volumes may be used at the
            convenience of the analyst.  If compound purity is certified
            at 96% or greater, the weight may be used without correction
            to calculate the concentration of the stock standard.
            Commercially prepared stock standards may be used at any
            concentration if they are certified by the manufacturer or by
            an independent source.
                               531.1-8

-------
           7.7.2  Transfer the stock standard solutions into TFE-fluoro-
                  carbon-sealed screw cap vials.   Store at room temperature  and
                  protect from light.

           7.7.3  Stock standard solutions should be replaced after two  months
                  or sooner if comparison with laboratory fortified blanks,  or
  -  • ,             QC samples indicate a  problem.

      7.8   INTERNAL  STANDARD S9LUTION - Prepare  an  internal  standard
           fortification  solution  by accurately weighing  approximately 0 0010 q
           of  pure BDMC.   Dissolve the BDMC  in methanol and  dilute  to volume in
           a 10-mL volumetric flask.   Transfer the internal  standard
           fortification  solution  to a TFE-fluorocarbon-sealed  screw cap bottle
           and  store at room temperature.  Addition  of  5  pi  of  the  internal
           standard  fortification  solution to  50  ml  of  sample results in a
           final  internal  standard concentration  of  10  /zg/L.  Solution should
           be replaced when  ongoing QC (Sect.  9)  indicates a  problem.  Note:
           BDMC  has  been  shown  to  be  an  effective internal standard for  the
           method  analytes,  but  other compounds may  be  used,  if the quality
           control requirements  in Sect.  9 are  met.

      7.9   LABORATORY PERFORMANCE  CHECK'SOLUTION  -  Prepare concentrate  by
           adding  20 pL of the 3-hydroxycarbofuran stock  standard solution,
           1.0 mL  of the  aldicarb  sulfoxide  stock standard solution, 200 #L  of
           the methiocarb stock  standard  solution, and  1  mL of  the  internal
           standard  fortification  solution to  a 10-mL volumetric flask.   Dilute
           to volume with methanol.   Thoroughly mix  concentrate.  Prepare check
           solution by placing 100 ./iL  of  the concentrate  solution into a 100-mL
           volumetric flask.  Dilute  to volume with  buffered  reagent water
           Transfer to a TFE-fluorocarbon-sealed  screw cap bottle and store  at
           room temperature.  Solution should be  replaced when  ongoing QC
           (Sect.  9)  indicates a problem.

8.   SAMPLE COLLECTION. PRESERVATION AND HANDLING

     8.1  Grab samples must be collected  in glass containers.  Conventional
          sampling practices (8)  should be followed; however, the bottle must
          not  be prerinsed with sample before collection.

     8.2  SAMPLE PRESERVATION/PH ADJUSTMENT - Oxamyl,  3-hydroxycarbofuran,
          aldicarb sulfoxide, and carbaryl can all  degrade quickly in  neutral
          and  basic  waters held at room temperature. (6,7)  This short  term
          degradation is  of concern during the time  samples  are being  shipped
          and  the time processed samples are held at room temperature  in
          autosampler trays.  Samples targeted .for the  analysis of these three
          analytes must be preserved at  pH 3.   The pH adjustment also
          minimizes  analyte biodegradation.

          8.2.1   Add 1.8  mL of monochloroacetic acid buffer  to  the  60-mL
                 sample bottle.   Add buffer  to the sample bottle at the
                 sampling site or in the laboratory  before shipping to the
                 sampling site.

                                   531.1-9

-------
          8.2.2  If residual chlorine is present, add 80 mg of sodium thio-
                 sulfate per liter of sample to the sample bottle prior to
                 collecting the sample.

          8.2.3  After sample is collected in bottle containing buffer, seal
                 the sample bottle and shake vigorously for 1 min.

          8.2.4  Samples must be iced or refrigerated at 4°C from the time of
                 collection until analysis is begun.  Although preservation
                 results of up to 28 days indicate method analytes are not
                 labile in water samples when sample pH is adjusted to 3 or
                 less, and samples are shipped and stored at 4°C, analyte
                 lability may be affected by the matrix.  Therefore, the
                 analyst should verify that the preservation technique is
                 applicable to the samples ;under study.

9.   QUALITY CONTROL

     9.1  Minimum quality control (QC) requirements are initial demonstration
          of laboratory capability, monitoring internal standard peak area or
          height in each sample and blank (when internal standard calibration
          procedures are being employed), analysis of laboratory reagent
          blanks, laboratory fortified samples, laboratory fortified blanks
          and QC samples.  A MDL for each analyte must also be determined.

     9.2  LABORATORY REAGENT BLANKS.  Before processing any samples, the
          analyst must demonstrate that all glassware and reagent
          interferences are under control.  Each time a set of samples is
          extracted or reagents are changed, a laboratory reagent blank (LRB)
          must be analyzed.  If within the retention time window of any
          analyte of interest the LRB produces a peak that would prevent the
          determination of that analyte, determine the source of contamination
          and eliminate the interference before processing samples.

     9.3  INITIAL DEMONSTRATION OF CAPABILITY.

          9.3.1  Select a representative concentration, about 10 times EDL, or
                 a concentration that represents a medium concentration
                 calibration standard for each analyte.  Prepare a primary
                 dilution standard (in methanol) containing each analyte at
                 1000 times selected concentration.  With a syringe, add 50 nl
                 of the concentrate to each of four to seven 50-mL aliquots of
                 reagent water, and analyze each aliquot according to
                 procedures beginning in Sect. 11.

          9.3.2  For each analyte the recovery value for all of these samples
                 must fall in the range of ± 20% of the fortified amount, and
                 the RSD of the measurements must be 20% or less.  For those
                 compounds that meet the acceptance criteria, performance is
                 judged acceptable and sample analysis may begin.  For those
                 compounds that fail these criteria, this procedure must be


                                   531.1-10

-------
            repeated using four samples until satisfactory performance
            has been demonstrated.

     9.3.3  To determine the MDL, analyze a minimum of 7 LFBs prepared at
           • a low concentration.  Use the concentrations in Table 3 as a
            guide, or use calibration data to estimate a concentration
            that will yield a peak with a signal to noise ratio of
            approximately 5.   Analyze the 7 replicates as described in
            Sect. 11,  and on a schedule that results in the analyses being
            conducted over several days.  Calculate the mean accuracy and
            standard  deviation for each analyte.  Calculate the MDL using
            the equation given in Table 3.

     9.3.4  The initial  demonstration of capability is used primarily to
            preclude  a laboratory from analyzing unknown samples via a
            new,  unfamiliar method prior to obtaining some experience
            with  it.   It is expected that as laboratory personnel  gain
            experience with this method the quality of data will improve
            beyond those required  here.

9.4  The a/?a,7yst  is permitted  to modify HPLC columns,  HPLC conditions,
     and internal  standards to improve separations  or lower analytical
     costs.  Each  time such method modifications are made,  the analyst
     mustirepeat the  procedures  in Sect.  9.3.

9.5  ASSESSING  THE INTERNAL STANDARD
9.5.1
            When using the  internal  standard  calibration  procedure,  the
            analyst must monitor the IS  response  (peak area  or peak.
            height) of all  samples during  each  analysis day.   The  IS
            response for any sample  chromatogram  should not  deviate  from
            the daily calibration check  standard's  IS  response by  more
                30%.
     9.5.2-If >30% deviation occurs with an  individual extract,  optimize
           [instrument performance and inject a second  aliquot.

           •9.5.2.1  If the reinjected aliquot produces an acceptable
                    internal standard response, report results for that
                    aliquot.
       K5.2.2
                    If a deviation of greater than 30% is obtained for
                    the reinjected extract, analysis of the sample
           ,         should be repeated beginning with Sect. 11, provided
           *'         a duplicate sample is still available.  Otherwise,
           •        report results obtained from the reinjected extract,
           V%       but annotate as suspect.

      9.5.3  If consecutive samples fail the IS response acceptance
            chterion,  immediately analyze a calibration check standard.
                             531.1-11

-------
            9.5.3.1  If the check standard provides a response for the IS
                     within 20% of the predicted value, then follow
                     procedures itemized in Sect. 9.5.2 for each sample
                     failing the IS response criterion.

            9.5.3.2  If the check standard provides a response for the IS
                     which deviates more than 20% of the predicted value,
                     then the analyst must recalibrate, as specified in
                     Sect. 10.                             •

9.6  ASSESSING LABORATORY PERFORMANCE - LABORATORY FORTIFIED BLANKS

     9.6.1  The laboratory must analyze at least one laboratory fortified
            blank (LFB) sample with every 20 samples or one per sample
            set (all samples analyzed within a 24-h period) whichever is
            greater.  The fortification concentration of each analyte in
            the LFB should be 10 times EI3L or a concentration in the
            middle of the calibration range.  Calculate accuracy as
            percent recovery (X{).   If the recovery of any analyte falls
            outside the control limits (see Sect. 9.6.2), that analyte is
            judged out'of control, and the source of the problem must be
            identified and resolved before continuing analyses.

     9.6.2  Until sufficient data become available from within their own
            laboratory, usually a minimum of results from 20 to 30
            analyses, the laboratory should assess laboratory performance
            against the control limits in Sect. 9.3.2.  When sufficient
            internal performance data becomes available, develop control
            limits from the mean percent recovery (X) and standard
            deviation (S) of the percent;recovery,  these data are used
            to establish upper and lower control limits as follows:

                        UPPER CONTROL LIMIT  = X + 3S
                        LOWER CONTROL LIMIT  = X - 3S

            After each five to ten new recovery measurements, new control
            limits should be calculated using only the most recent 20-30
            data points.  These calculated control limits should not
            exceed those established in Sect. 9:3.2.

     9.6.3  If acceptable accuracy and method detection limits cannot be
            achieved, the problem must be located and corrected  before
            further samples are analyzed,   Data from all field samples
            analyzed since the last acceptable LFB should be considered
            suspect, and duplicate samples should be analyzed, if they
            are available, after the problem has been corrected.  LFB
            results should be added to the on-going control charts to
            document data quality.

            Since the calibration check sample in Sect. 10.2.4 and 10.3.3
            and the LFB are made the same way and since procedural
                              531.1-12

-------
             standards are used, the sample analyzed here may also be used
             as a calibration check as described in those sections.

      9.6.4  It is recommended that the laboratory periodically determine
             and document its detection limit capabilities for analytes of
             interest.

      9.6.5  At least quarterly, analyze a QC sample from an outside
             source.

 9.7  ASSESSING ANALYTE RECOVERY - LABORATORY FORTIFIED SAMPLE MATRIX

      9.7.1  The laboratory must add a known concentration to a minimum of
             5% of the routine samples or one sample per set, whichever is
             greater.   The concentration should not be  less then the
             background concentration  of the sample selected for
             fortification.   Ideally,  the concentration should be the same
             as that  used for the laboratory fortified  blank (Sect.  9.6).
             Over  time,  samples  from all  routine sample sources should be
             fortified.

      9.7.2  Calculate the percent  recovery,  P,  of  the  concentration  for
             each  analyte,  after correcting  the analytical  result,1 X,  from
             the fortified sample for  the  background  concentration,  b,
             measured  in  the  unfortified  sample,  i.e.,:

             P  = 100  (X - b)  / fortifying  concentration,

             and compare  these values  to  control  limits  appropriate for
             reagent water data  collected  in  the  same fashion.   The value
             for P must fall  in  the  range  of  65%-135% of the  fortified
             amount.

     9.7.3   If  the recovery  of  any  such analyte  falls outside the
            designated range, and the laboratory performance for that
            analyte is shown to  be  in control  (Sect. 9.6), the  recovery
            problem encountered  with the dosed sample is judged to be
            matrix related,  not  system related.  The result  for that
            analyte in the unfortified sample  is labeled suspect/matrix
            to  inform the data user that the results are suspect due to
            matrix effects.

9.8  ASSESSING  INSTRUMENT SYSTEM - LABORATORY PERFORMANCE CHECK SAMPLE -
     Instrument performance should be monitored on a daily basis by
     analysis of the LPC sample.  The LPC sample contains compounds
     designed to monitor instrument sensitivity, column performance
     (primary column)  and chromatographic performance.   LPC sample
     components and performance  criteria are listed in  Table 4.
     Inability to  demonstrate acceptable instrument performance indicates
     the need for  revaluation of the  instrument system.  The sensitivity
     requirements  are  set based  on the EDLs  published in this method.  If
     laboratory EDLs  differ from those listed in this method,

                              531.1-.13

-------
          concentrations of the LPC standard compounds must be adjusted to be
          compatible with the laboratory ED|_s.

     9.9  The laboratory may adopt additional quality control  practices for
          use with this method.  The specific practices-that are most
          productive depend upon the needs of the laboratory and the nature of
          the samples.  For example, field or laboratory duplicates may be
          analyzed to assess the precision of the environmental  measurements
          or field reagent blanks may be used to assess contamination of
          samples under site conditions, transportation and storage.

10.  CALIBRATION AND STANDARDIZATION       ;

     10.1 Establish HPLC operating parameters equivalent to those indicated in
          Sect. 6.6.  The HPLC system may be calibrated using either the
          internal standard technique (Sect. 10.2) or the external standard
          technique (Sect. 10.3).

     10.2 INTERNAL STANDARD CALIBRATION PROCEDURE.  The analyst must select
          one or more internal standards similar in analytical behavior to the
          analytes of interest.  The analyst must further demonstrate that the
          measurement of the internal standard is not affected by method or
          matrix interferences.  BDMC has been identified as a suitable
          internal standard.

          10.2.1 Prepare calibration standards at a minimum of three
                 (recommend five) concentration levels for each analyte of
                 interest by adding volumes of one or more of the stock
                 standards to a volumetric flask.  Guidance on the number of
                 standards is as follows:  A minimum of three calibration
                 standards are required to calibrate a range of a factor of 20
                 in concentration.  For a factor of 50 use at least four
                 standards, and for a factor of 100 at least five standards.
                 The lowest standard should represent analyte concentrations
                 near, but above, their respective MDLs.  The remaining
                 standards should bracket the analyte concentrations expected
                 in the sample, or should define the working range of the
                 detector.  To each calibration standard, add a known constant
                 amount of one or more internal standards, and dilute to
                 volume with buffered reagent water.  To prepare buffered
                 reagent water, add 10 mL of 1.0 M monochlo.roacetic acid
                 buffer to 1 L of reagent water.

          10.2.2 Analyze each calibration standard according to the procedure
                 in Sect. 11.  Tabulate peak height or area responses against
                 concentration for each compound and internal standard.
                 Calculate response factors  (RF) for each analyte and
                 surrogate using Equation 1.
                    RF =
                             (A,)(CU)
Equation 1
                                   531.1-14

-------
               where:
                   As  = Response for the analyte to be measured
                Ais = Response for the  internal standard.
                Cjs = Concentration of  the  internal  standard /ig/L).
                Cs  = Concentration of the  analyte to be measured
      10.2.3 If the RF value over the working range is constant (20% RSD
             or less),  the average RF can be used for calculations
             Alternatively,  the results can be used to plot a calibration
             curve of response ratios (As/Ais) vs. Cs.

      10.2.4 The working  RF  or calibration curve must be verified on each
             working  day  by  the measurement of a minimum of two
             calibration  check standards,  one at the beginning and one at
             the end  of the  analysis  day.   These check standards  should be
             at two different  concentration levels  to verify the
             concentration curve.   For  extended  periods  of analysis
             (greater than 8 hr),  it  is  strongly recommended that check
             standards  be  interspersed  with samples  at regular intervals
             during the course of  the analyses.   If  the  response  for any
             analyte  varies from the  predicted response  by more than ±20%
             the  test must be  repeated  using  a fresh  calibration  standard'
            .If the results still  do  not  agree,  generate  a new calibration
             curve.

10.3  EXTERNAL  STANDARD CALIBRATION. PROCEDURE

      10.3.1  Prepare  calibration standards  as described  in  Sect 10.2  1
             omitting the use  of an internal  standard.

      10.3.2  Starting with the standard of  lowest concentration,  analyze
            each calibration  standard according to Sect.  11.2  and
            tabulate responses (peak height or area) versus the
            concentration in the standard.  The results can be used to
            prepare a calibration curve for each compound
            Alternatively, if the ratio of response to concentration
             (calibration  factor) is a constant over the working range
             (<20% relative standard deviation),  linearity through the
            origin can be assumed and the average ratio or calibration
            factor can be used in place of a calibration curve.

     10.3.3 The working calibration curve or calibration factor must be
            verified  on each  working  day as described in Sect. 10.2.4.

10.4 Verify calibration standards  periodically,  recommend at least
     quarterly  by analyzing  a standard prepared from reference material
     obtained from an independent  source.   Results  from these analyses
     must be within the limits used to  routinely check calibration (Sect
                              531.1-15

-------
11.  PROCEDURE

     11.1 PH ADJUSTMENT AND FILTRATION       :

          11.1.1 Add preservative to any samples (LFBs, LRBs, or calibration
                 standards) not previously preserved (Sect. 8).  Adjust the pH
                 of the sample or standard to pH 3 ± 0.2 by adding 1.5 ml of
                 2.5 M monochloroacetic acid buffer to each 50 ml of sample.
                 This step should not be necessary if sample pH was adjusted
                 during sample collection as a preservation precaution.  Fill
                 a 50-mL volumetric flask to the mark with the sample.  Add 5
                 jiL of the internal standard fortification solution (if the
                 internal standard calibration procedure is being employed)
                 and mix by inverting the flalsk several times.

          11.1.2 Affix the three-way valve to a 10-mL syringe.  Place a clean
                 filter in the filter holder and affix the filter holder and
                 the 7- to 10-cm syringe needle to the syringe valve.  Rinse
                 the needle and syringe with reagent water.  Prewet the filter
                 by passing 5 ml of reagent water through the  filter.  Empty
                 the syringe and check for leaks.  Draw 10 mL  of sample into
                 the syringe and expel through the filter.  Draw another 10 mL
                 of sample into the syringe, expel through the filter, and
                 collect the last 5 mL for analysis.  Rinse the syringe with
                 reagent water.  Discard the filter.                     •

     11.2 LIQUID CHROMATOGRAPHY

          11.2.1 Sect. 6.6 summarizes the recommended operating conditions  for
                 the liquid chromatograph.  Table 1 lists  retention times
                 observed  using this method.  Other HPLC columns or chromato-
                 graphic conditions may be used  if the requirements of Sect. 9
                 are met'.                    ;

          11.2.2 Calibrate or verify the system  calibration  daily  as  described
                 in Sect.  10.  The  standards and  samples must  be  in pH 3
                 buffered  water.        •

          11.2.3 Inject  400 /iL of  the sample.   Record  the  volume  injected  and
                 the resulting peak size in  area  units.

          11.2.4 If the  response for the peak exceeds  the  working  range  of the
                 system, dilute the sample with  pH 3  buffered  reagent water
                 and reanalyze.

     11.3 IDENTIFICATION OF ANALYTES

          11.3.1 Identify  a sample  component by  comparison of  its  retention
                 time  to the  retention  time  of  a reference chromatogram.   If
                 the retention time of  an  unknown compound corresponds,  within
                 limits, to the  retention  time  of a  standard compound, then
                 identification  is  considered  positive.

                                    531.1-16  ;

-------
           11.3.2 The width of the retention time window used to make
                 identifications should be based upon measurements of actual
                 retention time variations of standards over the course of a
                 day.  Three times the standard deviation of a retention time
                 can be used to calculate a suggested window size for a
                 compound.  However, the experience of the analyst should
                 weigh heavily in the interpretation of chromatograms.

           11.3.3 Identification requires expert judgement when sample
                 components are not resolved chromatographically.  When peaks
                 obviously represent more than one sample component (i.e.,
                 broadened peak with shoulder(s) or valley between two or more
                 maxima), or any time doubt exists over the identification of
                 a peak on a chromatogram, appropriate alternate techniques,
                 to help confirm peak identification, need to be employed.
               .  For example, more positive identification may be made by the
                 use of an alternative detector which operates on a
                 chemical/physical  principle different from that originally
                 used;  e.g.,  mass spectrometry (1), or the use of a second
                 chromatography column.   A suggested alternative column is
                 described in Sect.  6.6.3.
12.   CALCULATIONS
     Determine the concentration of individual  compounds in the sample using
     the following equation:

                     Cx    =    (AJ     (QJ
                              (As)     (RF>

          where:  Cx  =  analyte" concentration  in micrograms per  liter;
                 Ax  =  response of the  sample  analyte;
                 As  =  response of the  standard (either  internal or
                         external),  in units consistent with  those  used
                         for the analyte  response;

                 RF  =  response factor  (with  an external  standard, RF  = 1,
                      because the standard  is the  same  compound as  the
                      measured analyte; with  an  internal standard RF  is  a
                      unitless value);

                 Qs =  concentration of internal  standard present or
                         concentration of external  standard that produced As,
                         in  micrograms per liter.

    Use the multi-point  calibration established in Section 10 for  all
    calculations.   Do not use the daily  calibration verification data to
    quantitate analytes-in  samples.
                                  531.1-17

-------
13.  METHOD PERFORMANCE
     13.1 In a single laboratory, analyte recoveries from reagent water were
          used to determine analyte MDLs and EDLs and demonstrate method
          range.  Analyte recoveries and standard deviation about the percent
          recoveries at one concentration are given in Table 3.

     13.2 In a single laboratory, analyte recoveries from two standard
          synthetic ground waters were determined at one concentration level.
          Results were used to demonstrate applicability of the method to
          different ground water matrices.  Analyte recoveries from the two
          synthetic matrices are given in Table 2.
14.  REFERENCES
     3.
      4.
      5.
      6.
      7.
          Behymer, T.D., Bel
          of Benzidines and
          Liquid  Extraction
          Performance  Liquid
          in Methods for the
                  lar, T.A., Ho, J.S., Budde, W.L., "Determination
                  Nitrogen Containing Pesticides in Water by Liquid-
                  or Liquid-Solid Extraction and Reverse Phase High
                   Chromatography/Particle Beam/Mass Spectrometry"
                   Determination of Organic Compounds in Drinking
          Water. Supplement
          Exposure  Research
                  2 (1992).  EPA/600/R-92/129  USEPA, National
                  Laboratory, Cincinnati,Ohio 45268.
Moye, H.A., S.J. Sherrer, and P. A. St. John, "Dynamic Labeling of
Pesticides for High Performance Liquid Chromatography:  Detection of
N-Methylcarbamates and o-Phthalaldehyde," Anal. Lett. 10, 1049,
1977.

ASTM Annual Book of Standards, Part 11, Volume 11.02, D3694-82,
"Standard Practice for Preparation of Sample Containers and for
Preservation," American Society for Testing and Materials, Philadel-
phia, PA, p. 86, 1986.

"Carcinogens - Working with Carcinogens," Department of Health,
Education, and Welfare, Public Health Service, Center for Disease
Control, National Institute for Occupational Safety  and Health,
Publication No. 77-206, Aug.  1977.

"OSHA Safety and Health Standards, General  Industry," (29 CFR  1910),
Occupational Safety and Health Administration, OSHA  2206, (Revised,
January  1976).

"Safety  in Academic Chemistry Laboratories," American Chemical
Society  Publication,  Committee on  Chemical  Safety, 3rd Edition,
1979.

Foerst,  D. L.  and H.  A. Moye,  "Aldicarb  in  Drinking  Water via  Direct
Aqueous  Injection HPLC with Post  Column  Derivatization,"  Proceedings
of the  12th annual AWWA Water Quality Technology  Conference,  1984.
                                    531.1-18

-------
                                                                                  I
M1! iu i   i      Hollowen, and L. A. DalCortevo, "Determination of
N-Methylcarbamate Pesticides in Well Water by Liquid Chromatography
and Post Column Fluorescence Derivatization," Anal. Chem. 56, 2465
(1984).
ASTM Annual Book of Standards, Part 11, Volume 11.01, D3370-82
"Standard Practice for Sampling Water," American
and Materials, Philadelphia, PA, p. 130, 1986.
Society for Testing
                        531.1-19

-------
*17. TABLES. DIAGRAMS. FLOWCHARTS. AND VALIDATION DATA
                 TABLE 1.  RETENTION TIMES FOR METHOD ANALYTES


                                           Retention  Time(a)
                                              (minutes)

         Analvte	Primarv(1)   A1ternative(2)     Alternative'3*
     Aldicarb  sulfoxide        6.80
     Aldicarb  sulfone          7.77
     Oxamyl                     8.20
     Methomyl                   8.94
     3-Hydroxycarbofuran      13.65
     Aldicarb                  16.35
     Baygon                    18.86
     Carbofuran               19.17
     Carbaryl                  20.29
     Methiocarb               24.74
     BDMC                      25.28
15.0
15.2
17.4
18.4
23.
27,
29.
.3
,0
.3
29.,6
30.8
34.9
35.5
17.5
12.2
14.6
14.8
19
21.4
24.4
23.4
25.4
28.6
      
-------
a
CO
G£
LU
i

a

a£
-
i

^p
-
o
^^
^!
O

^j
***•
uu
CD
 O'
re
3
O
cu
t—
4J CC
E
00








•o
1 — 1

S-
cu
+J
re c
3 oo
o
i *
cu
-C

E
oo








a>o

re / s
— q» '

4-9

cu
cn
re.a
CU rv





E
O

CU 4->
•i— re
4- S 	 1
4-> E O
S- CU =3
O 0
U_ E
o.






•£
re
E
•=c



csjinrooo^Oi-H-^LOCM
(.ga^OLnoi^^csrocncD
i— 1 i— 1 i— 1 r-<



oScncrJooo^So














CT> CsJ OO h^. i— i CT^ r-H O^ CT\
f CXJ WJ U O i Ol




vo 00 LO to ^f* cxj 00 cvj 00 r*"*«.
r— 1 t-H r— i i-H









Lr>ocr»csjoO'-H'— icncvio
. . ^^ _ ^^ ^_^
• f* '^ "sj ^* ^^ "* "** *~* "**J *J"






^Swo^oSSSo










O O LO LO
LOOOLOOI — ^OCOCVJO
CSJ • '




( _
cu re
EX M-
^£ _§
i — i — S-
33 re
00 00 E U jQ
re >-, s-
-O-Q-C1 i — S_X«i —
i-s-s- ^soo^
rerereES-<-i-s-oEi —
15 ^ ^ ^-£ -£ .^5 ^ H
i — i — i — rerere i cucux
•=c«a:«a;cQc_3<_3co2:2icD


















•
oo
a>'
r—
d.
E
re
oo

OO
1
1--.

o

re
cu
E

cu
-E
"'""'
-*-*
E
CU
CU
. a.
O)
s_

"O
E
re

r
E
re
S
E
"p~
-a
cu
^— >
o
cu
-t-5
cu
-a
-1-9
E
O
E
re

S-
o
•a
cu
o
cu
S-
o
o
re
-t->
re
Q

a


























































•
>•>
S-
cu
o
o
a>
&.

•i-3
E
CU
S-
cu
Q.
CU
cn
re
cu
^
^
CU
>
o
o
cu
S—
E
CU
o
. s-
a>
a.

cu
-E
1 '

O
E
O
.^
-1-J
re
>r«
^
cu
T3
S-
re
E
re
00
ii
00*

U


^_
cu
re
3
~ 0)
3
a.
o

o
0

S-
cu
re
*^g
cn
E
"^
a.
00

E
re
'lo
cu
4-3

3 3
O O
re >,
s- 51
o
<+- E
cu
4-> >,
O E
cu re
s- a.
S- E
o o

•^




o
•r™
re
s- >>
E CU
CU •<-
U  CU
O 00
4-> -i- 3
« E
3 00
•o :n re
••- s
O i —
re re *»
o o o
> 4-9 re
i — re s_
3 E 0

CU O
-E 4-9 O
4-> E
*^ *— H r>
S S-
a> cu
•a -E >
CU 4-> E
-i- CU
<+- E Q
•r- O
-1-9 S- E
S— ^f -^—
o
M- CU >,
i — CU
S- -Q >
cu re s-
[ •* [ *— ;
re -i— oo
S re
> r—
4-9 re re
C (J
CU ^.-r-
cn-o cn
CU U i —
s- re o
cu
• " O C3
E > oo
re i — cu
i — 3 4->
("*i C^. 03
4^
E -a oo
•r- CU
isl -o
-a •!— cu
E S- 4->
3 CU ••-
O 4-> E
<+- 0 Z3
re
4-9 S- CU
E re .E
3 J= 4->
O U
E 1 .E
re i — 4-9
s- rcu 2
0 S
<+- -a
cu
cu re
0 • 0
CU i — O
S- CU 00
i- > 00
o cu re
O i — — •

01
531.1-21

-------
{2
o
a
s
a

S
co
LU

i
         S
         o  en
         ==  =3.
 1
 a
 O
 a

 S
 G
 Z LU

 SS
 loS


 Sz
 B£ LU
 O. CD

   «LU
 >* O£
 o
  < co
    LU
  O -J
  OS -*
  CO

  LU
          CU
          >
          o
          u
           O)
                  CVlCJi— I  CVJ  •— I CSJ
                      CTl
                                OJ




                             t-H O  i—<
                                          CT1  ID
                   CJ  CVJ CO CV4  >— 1 CM > — l
                 CO  CO CO  1 — OOf--COOOI~-00
                 OOOOOI^JOO


                    CVJ CM t—i CM i—I  CM *3-
                        QJ
                       -a
                    i  S-

                                  3 O  U  ^

                                  'o TO -2  o r>,
                                  J2 >> -C  -C  E
                                                       o
                                                       CU


                                                       E
•}->

(/)







E
O
"S
cu
S-
<4~ ,
t]
o

t/j
cu
cu
S-
O)
-o

1 — 1
1
c

-C

.^
3

r_
CU

CU


cu
o
c
cu
T3

<£
e
o
o

5^
en


cu

] •>

^_
o
t)

cu
3
^_
tO
>

-t-»
01
+J . (^
C CU
CU +J
"O tO
3 O
OO r^
Q.

^_
^•N
O* '1
• O
o
II S—

J2 _Q
3 E

(0 ^
v_
"" II
*^
C j£
^
+J





CO
rt cu
C i—
•r- (TJ
C T3
(O C S-
3 O)
C 0 >
o a. cu
._. & r™
•r— CS -^«
-t-> O O
•I- O •!-
C ^
i+^ 0
 >»
CU r—
ID i— CU
CO -M
^ "* E
4-* S- '^^
s- o x
rO O
CL. ^. S-
i— ( 0.
o; i— < ca.
u_ • to
<_j i—i
4-
0 C 0
•* 0
•r- O
O  -r- +->
> cO
CQ CU S-
Qi
x cu
•r- 1 01
-O •>-
c: -*J o
O) -i- C
a. E I
Q.-I- 0
^C ' -t-3
**-^ 1
C i—
_l O fO
Q "- C
S +-> cr>
01 0 -r-
CU S- CU 01
01 CU -!->
>, ^. CU -C
i— +J Q •*->
03 -i- -r-
C CU -O 3
tO O
01 JT +->
CU tO +-> O
+J ' CU tO
to -o s s-
o cu +J
•r- C CU X
r- -r- JC. 0)
r^ C|_ ^_J
cu cu . i—
o c
0 •»-» C <4-
•r- O
= E -r- 0)
O •!— +J -C
••- r- to -M
fO C -i— C
•i— O E •"—
> -r- S-
O) •*-> CU ^
TO O -4~* tO
CU CU CU
-o +-> Q a.
s- a>
ro -a cu to
-a -=
c TO -t-> en
to cu c
+J 4-> S_ -r-
(/> tO O "O
E <+-i—
II -r- CU
•4J CU -r-
OO 01 S- >i
CU 3
•o cu s-
ii cu i— cu
U CLJC
— 1 O E CD
Q i_ fO v-
UJ Q. 01 _G
                                                    531.1-22

-------


















z
o
|«H

J3
o
oo

^J
0
LLJ
Z
o
LU
O
Z
^
o£
o
u_
o:
UJ
o_

^B
^J*
O
1—
2
o
CD
_l


•
<3-
LU
_l
CD

•"—
c
CO
C£














"
c a
o -z.
o

















CO
•t"3
^
ra
£=

o
CO
4-J
CO
Q









cvj













ra
s-
4-
o
f">
S-
ro
O
O
s-
"5,
~T"
1
CO





'2

CU
GO












to
1 — 1
•
!-H
V
l_l_
Q-
V

0
o^
•
o









CD
0
I"' <












CO
•a
X
o
f 1
r__
3
V)
s-
rfl
o
-a
^—
^


CO
o
£=
ro
g
O
<4—
S-
co
CL
O
CL
ra
en'
o
ro
E
O
.c •
<_)











-Q
0
t— i
A
C
O
•r—
^_J
3
^—
0
oo
0)
a:









CD CD
O\J i— I






^~
>-j
£= ;
CO
r" f^~^
O-GO
r— i— <
-C
+-> CO
CO -l->
E ro
•r- E
~O ra

ur> s-
XJ - ro
s- ro o
ro 1 r—
CJ O >>
O E -C
-r- O 4->
J= S- CO
-(-> 03 E
CU 1 1
SI -=t- 21





performance
c
3
r—
O
o

































c
o
•r-
^^
ro
3 •
cr
cu

cu
-£=
4-J
cp
00
^t

"O
CO

ro
3
O
'co



•
^
O -^
O ^~
ro •— '
, 3
r5 X
3
ro •— '
0 II
jj<:
ro 4:
0) O
CL °-
u_
0
Q.

to

-C

2^
O)
-(J
-I-J
ro

•
00
CO
00
c

^r
•o
3

sX
ro
CO


CO
+J

oo
•i—

^-^
O
i — i
i — i
3 '
-a
c
ra

.
oo
(J
CO
oo

c
•r—
£
en
CO
<—

1 1
t
ra
ro
_£T

-^j
-r—
3
^^ ro
r— 1
i~H ^^
3
00
CVJ
3 •
CO -C
s- en
CO -r-
^= cu


































"
o
•r—
4~>
ra
3
cr .
CO

CO
r"
-1_>

Jd
CO

.^_
f I
CO

oo
ra
00
ra
CO
CL

o
3
cu
+->
,_
co
CO
CO
J2
c
o
+->
1 ^

^
rt
n
cu
OO

.c


4J
-a

3
ra
CO
CL
CO
en
ro
5-
cu

ra

cu

1 *

00
"'
3

T3

ra

00
ra
CO
CL
o
3


CO

J t

c:
cu
0)
3
CO
oo
CO

^'
. \
'
1=
o
•f-"
^
CU. 00

c: ra
•i— cu
CL
cu
(J O
CO 4->
co cu
uZ 5
-O <4-
o
cu
4-> CU
c
00 •!—
O)
_. 4-» 00
3 ra
CO JD
s-
QJ CU
3 -!->


531.1-23

-------
THIS PAGE LEFT BLANK INTENTIONALLY
             531.1-24

-------
      METHOD  551.1
DETERMINATION OF CHLORINATION DISINFECTION BYPRODUCTS
CHLORINATED SOLVENTS, AND HALOGENATED PESTICIDES/
HERBICIDES IN DRINKING WATER BY LIQUID-LIQUID
EXTRACTION AND GAS CHROMATOGRAPHY WITH ELECTRON-
CAPTURE DETECTION
                                 Revision  1.0
J.W. Hodgeson, A.L. Cohen - Method 551, (1990)

D.J. Munch (USEPA, Office of Water) and D.P. Hautman (International
           Consultants, Inc.) - Method 551.1, (1995)
                    NATIONAL  EXPOSURE RESEARCH LABORATORY
                      OFFICE OF RESEARCH AND DEVELOPMENT
                     U.S.  ENVIRONMENTAL  PROTECTION AGENCY
                            CINCINNATI,  OHIO 45268
                                   551.1-1

-------
                                 METHOD 551.1

 DETERMINATION OF CHLORINATION DISINFECTION BYPRODUCTS,  CHLORINATED SOLVENTS,
   AND HALOGENATED PESTICIDES/HERBICIDES IN DRINKING WATER BY LIQUID-LIQUID
       EXTRACTION  AND  GAS  CHROMATOGRAPHY WITH  ELECTRON-CAPTURE DETECTION


1.   SCOPE AND APPLICATION

     1.1  This method (1-9) is applicable to the determination of the
          following analytes in finished drinking water, drinking water during
          intermediate stages of treatment, and raw source water.  The
          particular choice of analytes from this list should be a function of
          the specific project requirements.

          Disinfection Byproducts (DBPs);
                             Analvte                         CAS No.
          Trihalomethanes    Chloroform                      67-66-3
                             Bromodichloromethane            75-27-4
                             Bromoform                       75-25-2
                             Dibromochloromethane           124-48-1
          Haloacetonitriles  Bromochloroacetonitrile      83463-62-1
                             Dibromoacetonitrile           3252-43-5
                             Dichloroacetonitrile          3018-12-0
                             Trichloroacetonitrile          545-06-2
          Other DBFs         Chloral  Hydrate                 75-87-6
                             Chloropicrin                    76-06-2
                             l,l-Dichloro-2-propanone       513-88-2
                             l,l,l-Trichloro-2-             918-00-3
                             propanone
          Chlorinated Solvents:
                             Carbon Tetrachloride            56-23-5
                             l,2-Dibromo-3-;                 96-12-8
                             chloropropane [DBCP]
                             1,2-Dibromoethane  [EDB]        106-93-4
                             Tetrachloroethylene            127-18-4
                             1,1,1-Trichloroethane           71-55-6
                             1,1,2-Trichloroethane           79-00-5
                             Trichloroethylene               79-01-6
                             1,2,3-Trichloropropane          96-18-4
          Pesti ci des/Herbi ci des;
                             Alachlor                    15972-60-8
                             Atrazine                     1912-24-9
                             Bromacil                       314-40-9
                             Cyanazine                    21725-46-2
                             Endrin                          72-20-8
                             Endrin Aldehyde               7421-93-4
                             Endrin Ketone               53494-70-5
                             Heptachlor                      76-44-8
                             Heptachlor  Epoxide            1024-57-3
                             Hexachlorobenzene              118-74-1
                                    551.1-2

-------
                              Hexachlorocyclopentadiene       77-47-4
                              Lindane  (gamma-BHC)             58-89-9
                              Metolachlor                  51218-45-2
                              Metribuzin                   21087-64-9
                              Methoxychlor                    72-43-5
                              Trifluralin                   1582-09-8

     1.2  This analyte list includes twelve commonly observed chlorination-
          disinfection byproducts (3,4), eight commonly used chlorinated
          organic solvents and sixteen halogenated pesticides and herbicides.

     1.3  This method is intended as a stand-alone procedure for either the
          analysis of only the trihalomethanes (THMs) or for all the
          chlorination disinfection by-products (DBPs) with the chlorinated
          organic solvents or as a procedure for the total analyte list.  The
          dechlorination/preservation technique presented in section 8 details
          two different dechlorinating agents.  Results for the THMs and the
          eight solvents may be obtained from the analysis of samples
          employing either dechlorinating agent. (Sect. 8.1.2)

     1.4  After an analyte has been identified and quantitated in an unknown
          sample with the. primary GC column (Sect. 6.9.2.1) qualitative
          confirmation of results is strongly recommended by gas
          chromatography/mass spectrometry (GC/MS) (10,11), or by GC analysis
          using a dissimilar column (Sect.  6,9.2.2).

     1.5  The experimentally determined method detection limits (MDLs) (12)
          for the above listed analytes are provided in Tables 2 and 8.
          Actual MDL values will vary according to the particular matrix
          analyzed and the specific instrumentation employed.

     1.6  This method is restricted to use by or under the supervision of
          analysts experienced in the use of GC and in the interpretation of
          gas chromatograms.  Each analyst must demonstrate the ability to
          generate acceptable results with this method-using the procedure
          described in Sect. 9.4.

     1.7  Methyl-t-butyl ether (MTBE) is recommended as the primary extraction
          solvent in this method since it effectively extracts all of the
          target analytes 1isted in Sect.  1.1.  However, due to safety
          concerns associated with MTBE and the current use of pentane by some
          laboratories for certain method analytes, pentane is offered as an
          optional extraction solvent for all  analytes except chloral hydrate.
          If project requirements specify the analysis of chloral hydrate,
          MTBE must be used as the extracting solvent.  This method includes
          sections specific for pentane as an .optional solvent.

2.   SUMMARY OF METHOD

     2.1  A 50 ml sample aliquot is extracted with 3 ml of MTBE or 5 mL of
          pentane.  Two /*L of the extract is then injected into a GC equipped


                                    551.1-3

-------
          with a fused silica capillary column and linearized electron capture
          detector for separation and analysis.  Procedural standard
          calibration is used to quantitate;method analytes.

     2.2  A typical sample can be extracted and analyzed by this method in 50
          min for the chlorination by-products/chlorinated solve'nts and 2 hrs.
          for the total analyte list.  Confirmation of the eluted compounds
          may be obtained using a dissimilar column (6.9.2.2) or by the use of
          GC-MS.  Simultaneous confirmation,can be performed using dual
          primary/confirmation columns installed in a single injection port
          (Sect. 6.9.3) or a separate confirmation analysis.

3.   DEFINITIONS

     3.1  INTERNAL STANDARD (IS) -- A pure analyte(s) added to a sample,
          extract, or standard solution in known amount(s) and used to measure
          the relative responses of other method analytes and surrogates that
          are components of the same sample or solution.  The internal
          standard must be an analyte that ;is not a sample component.

     3.2  SURROGATE ANALYTE (SA) — A pure analyte(s), which is extremely
          unlikely to be found in any sample, and which is added directly to a
          sample aliquot in known amount(s) before extraction or other
          processing and is measured with the same procedures used to measure
          other sample components. The purpose of a surrogate analyte is to
          monitor method performance with each sample.

     3.3  LABORATORY DUPLICATES (LD1 and LD2) — Two sample aliquots, taken in
          the laboratory from a single samp'Je bottle, and analyzed separately
          with identical procedures.  Analyses of LD1 and LD2 indicate
          precision associated with laboratory procedures, but not with sample
          collection, preservation, or storage procedures.   This method
          cannot utilize laboratory duplicates since sample extraction must
          occur in the sample vial and sample transfer is not possible due to
          analyte volatility.

     3.4  FIELD DUPLICATES (FD1 and FD2) — Two separate samples collected at
          the same time and place under identical circumstances and treated
          exactly the same throughout field and laboratory procedures.
          Analyses of FD1 and FD2 give a measure of the precision associated
          with sample collection, preservation and storage, as well as with
          laboratory procedures.  Since laboratory,duplicates cannot be
          analyzed, the collection and analysis of field duplicates for this
          method is critical.

     3.5  LABORATORY REAGENT BLANK (LRB) —An aliquot of reagent water, or
          other blank matrix, that is treated exactly as a sample including
          exposure to all glassware, equipment, solvents, reagents, internal
          standards, and surrogates that are used with other samples.  The LRB
          is used to determine if method analytes or other interferences are
          present in the laboratory environment, the reagents, or the
          apparatus.

                                    551.1-4:

-------
3.6   FIELD  REAGENT  BLANK  (FRB)  —  Reagent water, or other  blank matrix,
      that is  placed in  a  sample container in  the laboratory  and treated
      as  a sample  in all respects,  including shipment to  sampling site,
      exposure to  sampling  site  conditions, storage, preservation and all
      analytical procedures.  The purpose of the FRB is to  determine if
      method analytes  or other interferences are present  in the field
      environment.

3.7   LABORATORY FORTIFIED  BLANK (LFB) — An aliquot of reagent water, or
      other  blank  matrix,  to which  known quantities of the  method analytes
      are added in the laboratory.  The LFB is analyzed exactly like a
      sample,  and  its, purpose is  to determine whether the methodology is
      in  control,  and  whether the laboratory is capable of  making accurate
      and precise  analyte  quantitation at various concentrations including
      the required method  detection limit.

3.8   LABORATORY FORTIFIED  SAMPLE MATRIX (LFM) — An aliquot  of an
      environmental  sample  to which known quantities of the method
      analytes  are added in the  laboratory.  The LFM is analyzed exactly
      like a sample,  and its purpose is to determine whether  the sample
      matrix contributes bias to  the analytical results. The  background
      concentrations  of the analytes in the sample matrix must be
      determined in  a  separate aliquot and the measured values in the LFM
      corrected for  background concentrations.

3.9   STOCK STANDARD  SOLUTION (SSS) — A concentrated solution containing
      one or more method analytes which is prepared in the  laboratory
      using assayed  reference materials or purchased as certified from a
      reputable commercial  source.

3.10  PRIMARY  DILUTION STANDARD  SOLUTION (PDS) — A solution  of several '
      analytes prepared in  the laboratory from stock standard solutions
      and diluted as needed to prepare calibration solutions  and other
      needed analyte solutions.

3.11  CALIBRATION STANDARD  (CAL)— A solution prepared from  the primary
      dilution standard solution  or stock standard solutions  and the.
      internal  standard(s)  and surrogate analyte(s).  The CAL solutions
      are used to calibrate the  instrument response with respect to
      analyte concentration.

3.12  QUALITY CONTROL SAMPLE (QCS) — A solution  of method analytes which
      is used to fortify an aliquot of LRB or sample matrix.  The QCS is
      obtained from a source external  to the laboratory and different from
      the source of calibration  standards.   It is used to check laboratory
      performance with externally prepared test materials.

3.13  LABORATORY PERFORMANCE CHECK SOLUTION (LPC)  -- A solution of
      selected method analytes,  surrogate(s),  internal  standard(s),  or
     other test substances used to evaluate the  performance of the
      instrument system with respect to a defined set of method criteria.
                               551.1-5

-------
     3.14 METHOD DETECTION LIMIT (MDL) — The minimum concentration of an
          analyte that can be identified, measured and reported with 99%
          confidence that the analyte concentration is greater than zero.
          (Appendix B to 40 CFR Part 136)

     3.15 ESTIMATED DETECTION LIMIT (EDL) — Defined as either the MDL or a
          level of compound in a sample yielding a peak in the final extract
          with a signal to noise (S/N) ratio of approximately 5, whichever is
          greater.

     3.16 PROCEDURAL STANDARD CALIBRATION —  A calibration method where
          aqueous calibration standards are prepared and processed (e.g.
          purged,extracted, and/or deri'vatized) in exactly the same manner as
          a sample..  All steps in the process from addition of sampling
          preservatives through instrumental, analyses are included in the
          calibration.  Using procedural standard calibration compensates for
          any inefficiencies in the processing procedure.

4.   INTERFERENCES

     4.1  Impurities contained in the extracting solvent usually account for
          the majority of the analytical problems.  Each new bottle of solvent
          should be analyzed for interferences before use,.  An interference
          free solvent is a solvent containing no peaks yielding data at > MDL
          (Tables 2 and 8) at the retention times of the analytes of interest.
          Indirect daily checks on the extracting solvent are obtained by
          monitoring the laboratory reagent blanks (Sect. 9.3).  Whenever an
          interference is noted in the reagent blank, the analyst should
          analyze the solvent separately to determine if the source of the
          problem is the solvent or another reagent.

     4.2  Glassware must be scrupulously cleaned (13).  Clean all glassware as
          soon as possible after use by thoroughly rinsing with the last
          solvent used in it.  Follow by washing with hot water and detergent
          and thoroughly rinsing with tap and reagent water.  Drain dry, and
          heat in an oven or muffle furnace at 400°C for 1 hr.   Do not muffle
          volumetric ware but instead rinse three times with .HPLC grade or
          better acetone.  Thoroughly rinsing all glassware with HPLC grade or
          better acetone may be substituted for heating provided method blank
          analysis confirms no background interferant contamination is
          present.  After drying and cooling, seal and store all glassware in
          a clean environment free of all potential contamination.  To prevent
          any accumulation of dust or other contaminants, store glassware
          inverted on clean aluminum foil or capped with aluminum foil.

     4.3  Commercial lots of the extraction solvents (both MTBE and pentane)
          often contain observable amounts of chlorinated solvent impurities,
          e.g., chloroform, trichloroethylene, carbon tetrachloride.  When
          present, these impurities can normally be removed by double
          distillation.
                                    551.1-6

-------
     4.4
This liquid/liquid extraction technique efficiently extracts a
            3 of non-polar and polar organic components of the
            ;. rnnfirmat.inn i <: miifp imnnvtant  navMrnl avO w =
wide
     4.5
     4.6
This liquid/liquid extraction technique efficiently extracts a w
boiling range of non-polar and polar organic components of the
sample.  Thus, confirmation  is quite important, particularly at
lower analyte concentrations.  A confirmatory column (6.9.2.2) is
suggested for this purpose.  Alternatively, GC/MS may also be used
for confirmation if sufficient concentration of analyte is present.

Special care must be taken in the determination of endrin since it
has been reported to breakdown to aldo and keto derivatives upon
reaction with active sites in the injection port sleeve (14).  The
active sites are.usually the result of micro fragments of the
injector port septa and high boiling sample residual deposited in
the injection port sleeve or on the inner wall at the front of the
capillary column.  The degradation of endrin is monitored using the
Laboratory Performance Check Standard (Sect. 9.2).

Interfering and erratic peaks have been observed in method blanks
within the retention windows of metribuzin, alachlor, cyanazine and
heptachlor.  These are believed to be due to phthalate
contamination.  This contamination can be reduced by paying special
attention to reagent preparation (See solvent rinsing the dry buffer
and the dechlorination/ preservative salts, Sect. 7.1.7.5) and
elimination of all  forms of plastic from the procedure (i.e.  HOPE
bottles,  plastic weighing boats, etc.).   After NaCl or Na2S04  is
muffled or phosphate buffer and dechlorination/preservative salts
are solvent rinsed,  they should be stored in sealed glass
containers.  NaCl,  Na2S04, phosphate buffer and dechlorination/
preservative salts should be weighed using glass beakers,  never
plastic weighing boats.
5.   SAFETY
     5.1  The toxicity and carcinogenicity of chemicals used in this method
          have not been precisely defined; each chemical should be treated as
          a potential health hazard, and exposure to these chemicals should be
          minimized.  Each laboratory is responsible for maintaining awareness
          of OSHA regulations regarding safe handling of chemicals 'used in
          this method.   Additional references to laboratory safety are
          available (15-17) for the information of the analyst.

     5.2  The following have been tentatively classified as known or suspected
          human or mammalian carcinogens:

          Chloroform, l,2-Dibromo-3-chloropropane, 1,2-Dibromoethane,
          heptachlor, and hexachlorobenzene.

     5.3  The toxicity of the extraction solvent, MTBE, has not been well
          defined.  Susceptible individuals may experience adverse affects
          upon skin contact or inhalation of vapors.   Therefore,  protective
          clothing and gloves should be used and MTBE should be used only  in a
          chemical fume hood or glove box.  The same  precaution applies to
          pure standard materials.
                                   551.1-7

-------
6.   EQUIPMENT AND SUPPLIES  (All specifications in Sections 6 and 7 are
     suggested.  Catalogue numbers are included for illustration only.)

     6.1  SAMPLE CONTAINERS - 60-mL screw cap glass vials (Kimble #60958A-16,
          Fisher #03-339-5E or equivalent) each equipped with size 24-400 cap
          and PTFE-faced septa (Kimble #738,02-24400, Fisher #03-340-14A or
          equivalent)r Prior to use or following each use,  wash vials and
          septa with detergent and tap water then rinse thoroughly with
          distilled water.  Allow the vials: and septa to dry at room
          temperature, place only the vials in an oven and heat to 400°C for
          30 min.  After removal  from the oven allow the vials to cool in an
          area known to be free of organics.  After rinsing caps with
          distilled water, rinse in a beaker with HPLC grade or better acetone
          and place in a drying oven at 80°C for 1  hr.

     6.2  VIALS - Autosampler,  2.0-mL vial with screw or crimp cap and a
          teflon faced septa.              :

     6.3  MICRO SYRINGES - 10 pi, 25 pi, 50 pi, 100 pi, 250 pi, and 1000 fjl.

     6.4  PIPETTES - 3.0 mL or 5.0 mL, type A, TD,  glass.

     6.5  VOLUMETRIC FLASK - 10 mL, 100 mL, 250 mL, 500 mL glass stoppered.

     6.6  DISPOSABLE PASTEUR PIPETS - 9 inch, used,for extract transfer.

     6.7  STANDARD SOLUTION STORAGE CONTAINERS - 30-mL Boston round, amber
          glass bottles with TFE-lined caps or equivalent.

     6.8  BALANCE - Analytical, capable of accurately weighing to the nearest
          0.0001 g.                        :

     6.9  GAS CHROMATOGRAPHY SYSTEM        '     ,   -

          6.9.1  The GC must be capable of temperature programming and should
                 be equipped with a linearized electron capture detector
                 (ECD), fused silica capillary column,  and on-column or
                 splitless injector (splitless mode, 30 sec. delay).  If
                •simultaneous confirmation is employed the GC must have a
                 second ECD. An auto-sampler/injector is desirable.

                 6.9.1.1  SPECIAL PRECAUTION:  During method development, a
                          problem was encountered with the syringe on the
                          autosampler.  The syringe plunger, after repeated
                          sample extract injections, developed resistance when
                          withdrawn or inserted into the syringe barrel.  This
                          resistance was dufe to salt deposits in the syringe
                          barrel  which were left behind following the
                          evaporation of hydrated MTBE. To minimize this
                          problem, a unique' syringe wash procedure was
                          employed.  After sample injection, the syringe was
                          first rinsed three times  with reagent water then

                                    551.1-81

-------
                three times with MTBE.  This effectively removed all
                the residual salt after each injection from the
                syringe and surmounted the problem.  Some
                autosampler designs may not encounter this problem
                but this precaution has been mentioned to alert the
                analyst. When pentane was used as the extraction
                solvent, this was not a problem.

6.9.2  Two GC columns are recommended.  Column A is recommended as
       the primary analytical column unless routinely occurring
       analytes are not adequately resolved.  Column B is
       recommended for use as a confirmatory column when GC/MS
       confirmation is not sensitive enough or unavailable.  Other
       GC columns or conditions may be employed provided adequate
       analyte resolution is attained and all the quality assurance
       criteria established in Sect. 9 are met.

       6.9.2.1  Column A - 0.25 mm ID x 30 m fused silica capillary
                with chemically bonded methyl polysiloxane phase
                (J&W, DB-1,-1.0 m film thickness or equivalent).  As
                a result of the different boiling points of MTBE
                (b.p. 55°C)  and pentane (b.p.  35°C),  two  different
                GC oven temperature programs are specified in Table
                1 for MTBE and Table 12 for pentane.  Retention
                times for target analytes were slightly different
                using the pentane oven temperature program but
                elution order, analyte resolution, and total
                analysis time were not effected.  Injector
                temperature:  200°C equipped with 4 mm ID
                deactivated sleeve with wool plug (Restek #20781 for
                HP GC's or equivalent).  This sleeve design was.
                found to give better analyte response than the
                standard 2 mm sleeve.  Detector temperature: 290°C.

       6.9.2.2  Column B - 0.25 mm ID x 30 m with chemically bonded
                6 % cyanopropylphenyl/94 % dimethyl polysiloxane
                phase (Restek, Rtx-1301, 1.0 /im film thickness or
                equivalent).  The column oven was temperature
                programmed exactly as indicated for column A (Tables
                1 and 12).  Injector and detector temperatures at
                200°C and 290°C,  respectively.    The  same
                temperature program was utilized to allow for
                simultaneous confirmation analysis.

6.9.3  To perform simultaneous confirmation from a single injection
       onto both the primary and confirmation columns, two injector
       setups can be employed.

       6.9.3.1  Using a two hole graphite ferrule (Restek #20235, or
                equivalent) both columns can be inserted into one
                injection port.  To ensure the column ends are
                centered in the injection port sleeve and not angled

                          551.1-9

-------
                          to the  side,
                          at the  base
                          equivalent).
                          this manner
                          the base of
                          twisting as
                          minimize thi
                          twisted four
                          seated.
 an inlet adaptor fitting is installed
of the injection port (Restek #20633,  or
  Use caution when installing columns  in
to ensure the column does not break at
the injector due to the two columns
the ferrule nut is tightened.  To
s hazard, the ferrule nut can be reverse
 to five times once the ferrule has been
                 6.9.3.2  An alternate procedure  involves  installing a  1 meter
                          portion of 0.25 mm deactivated,  uncoated fused
                          silica capillary tubing  (Restek  #10043, or
                          equivalent) into the  injector as a normal single
                          column is installed.  Then using a Y-press tight
                          union (Restek #20403  or  equivalent) join the  1 meter
                          uncoated column to the  primary and secondary
                          columns.  Using this  procedure will reduce detection
                          limits when compared  to  the procedure outline in
                          6.9.3.1 since only one  column is positioned in the  .
                          injection port to receive the injected sample
                          extract.

          6.9.4  The analyst is permitted to modify GC columns, GC conditions,
                 internal standard or surrogate compounds.  Each time such
                 method modifications are made, the analyst must repeat the
                 procedures in Sect. 9.4.  '•

7.   REAGENTS AND STANDARDS

     7.1  REAGENTS

          7.1.1  MTBE - High purity grade. It may  be necessary to double
                 distill the solvent if impurities are observed which coelute
                 with some of the more volatile compounds.

          7.1.2  Pentane (optional extraction solvent) - High purity grade. It
                 may be necessary to double distill the solvent if impurities
                 are observed which coelute with some of the more volatile
                 compounds.

          7.1.3  Acetone - High purity,  demonstrated to be free of analytes.

          7.1.4  Methanol - High purity, demonstrated to be free of analytes.

          7.1.5  Sodium Chloride,  NaCl  - ACS Reagent Grade.  Before using a
                 batch of NaCl, place in muffle furnace, increase temperature
                 to 400°C and  hold  for 30 rtrin.   Store  in a capped glass
                 bottle, not in a plastic container.

          7.1.6  Sodium Sulfate, Na2S04 - ACS Reagent Grade.  Before using  a
                 batch of Na2S04, place  in muffle  furnace,  increase
                                   551.1-10

-------
       temperature to 400°C and hold for 30 min.   Store in a capped
       glass bottle not in a plastic container.

7.1.7  Sample Preservation Reagents

       7.1.7.1  Phosphate buffer - Used to lower the sample matrix
                pH to 5.2 in order to inhibit base catalyzed
                degradation of the haloacetonitriles (7), some of
                tfye chlorinated solvents, and to standardize the pH
                of all samples.  Prepare a dry homogeneous mixture
                of 2.50% Sodium Phosphate, dibasic (Na2HP04)/97.5%
                Potassium Phosphate, monobasic (KH2P04) by weight
                (example:   5.00 g Na2HPO,  and  195 g KH2PO, to yield a
                total  weight of 200 g)   Both of these buffer salts
                should be in granular form and of ACS grade or
                better.   Powder would be ideal but would  require
                extended cleanup time as outlined below in Sect.
                7.1.7.5  to allow for buffer/solvent settling.

       7.1.7.2  Ammonium Chloride,  NH^Cl,  ACS  Reagent  Grade.   Used
                to convert free chlorine to monochloramine.
                Although this is not the traditional  dechlorination
                mechanism, ammonium chloride is categorized as a
                dechlorinating agent in this method.

       7.1.7.3  Sodium Salfite, Na2S03,  ACS Reagent Grade.  Used as
                a  dechlorinating agent  for chloral  hydrate sample
                analysis.

       7.1.7.4  To simplify the addition of 6.0 mg  of the
                dechlorinating agent to the 60 mL vial, the
                dechlorinating salt is  combined with  the  phosphate
                buffer as  a homogeneous mixture.  Two  g of the
                appropriate dechlorinating agent  are  added to 200  g
                of the phosphate buffer.   When 0.60 g  of  the
                buffer/dechlorinating agent mixture are added to the
                60-mL  sample  vial,  6 mg of the dechlorinating agent
                are included  reflecting an actual concentration  of
                100 mg/L.   Two separate mixtures  are  prepared, one
                containing ammonium chloride  and  the other with
                sodium sulfite.

       7.1.7.5   If background  contaminants are detected in the salts
                listed in  Sections  7.1.7.1 through 7.1.7.3,  a
                solvent  rinse  cleanup procedure may be required.
                These  contaminants  may  coelute with some  of the  high
                molecular  weight  herbicides  and pesticides.   These
                salts  cannot  be  muffled  since  they decompose  when
                heated to  400°C.  This solvent rinsing procedure is
                applied  to  the  homogeneous mixture prepared in Sect.
                7.1.7.4.
                        551.1-11

-------
7.2
7.3
                 NOTE:  If a laboratory is not conducting analyses
                 for the-high molecular weight herbicides and
                 pesticides, this;cleanup may not be required if no
                 interfering peaks are observed within the retention
                 time window (Sect.12.2) for any target analytes in
                 the laboratory reagent blank.

                 SOLVENT RINSE CLEANUP PROCEDURE

                 Prepare two separate homogeneous mixtures of the
                 phosphate buffer ^alts (Sect. 7.1.7.1) in a 500-mL
                 beaker.  To one, add the correct amount of ammonium
                 chloride and to the other add the correct amount of
                 sodium sulfite.  Three separate solvents are then
                 used to rinse the mixture.   (This solvent rinsing
                 must be performed in a fume hood or glove box.)
                 First,  add approx. 100 mL of methanol, or enough to
                 cover the salt to a depth of approx.  1 cm,  and  using
                 a  clean spatula, stir the solvent salt mixture  for 1
                 minute.  Allow the buffer/solvent mixture to settle
                 for 1 minute and then decant the methanol,  being
                 careful not to pour off the rinsed buffer.   It  may
                 be necessary to perform this procedure up to four
                 times with methane!.   NOTE:  By softly  lifting and
                 tapping the base of the beaker against the  fume hood
                 counter surface,  more of the solvent  is brought to
                 the surface of the buffer.   Next,  perform the
                 identical  procedure up to two times using acetone.
                 Finally,  perform two  final  rinses  with the
                 appropriate extracting solvent (MTBE or Pentane).
                 After the  final  solvent rinse,  place the "wet"
                 buffer  on  a hot plate at  approx.  60°C  for 30 minutes
                 or until  dry.   Stir the mixture  every  5 minutes to
                 aid the evaporation  of excess solvent.   Once dry,
                 place the  buffer in  a glass  bottle  with either  a
                 ground  glass  stopper  or TFE  faced  septum.

REAGENT WATER -  Reagent water  is  defined  as  purified water  which
does not contain any  measurable  quantities  of any  target  analytes  or
any other interfering  species.

7.2.1  A Millipore  Super-Q water  system or  its equivalent may be
       used to generate deionized  reagent water.  Distilled  water
       that has  been  charcoal  filtered  may  also  be  su.itable.

7.2.2  Test reagent water each  day  it  is  used  by analyzing  according
       to Sect.  11.

STOCK STANDARD SOLUTIONS (SSS)- These  solutions may be  obtained  as
certified solutions or  prepared from  neat materials using the
following procedures:
                              551.1-12

-------
7.3.1   For  analytes which  are  solids  in their  pure  form, prepare
       ,stock  standard  solutions  (1 mg/mL)  by accurately weighing
        approximately 0.01  g of pure material in  a 10-mL volumetric
        flask.  Dilute  to volume  with  acetone.  Due  to the low
        solubility of simazine, this stock  should be prepared at 0.5
        mg/mL  by weighing 0.005 g diluted to volume with acetone in a
        10-mL  volumetric flask.   Alternatively, simazine stock
        standard solutions  may  be prepared  in ethyl  acetate at
        approximately 0.01  g/10 ml.  Stock  standard  solutions for
        analytes which  are  liquid in their  pure form at room
        temperature can be  accurately  prepared  in the following
        manner.

        7.3.1.1  Place  about 9.8  mL of acetone  into  a 10-mL ground-
                glass  stoppered  volumetric flask.  Allow the flask
                to stand,  unstoppered, for about 10 min to allow
                solvent film to  evaporate  from the  inner walls of
                the volumetric flask, and weigh to the nearest 0.1
                mg.

     1   7.3.1.2  Use a  10-/iL syringe and immediately add 10.0//L of
                standard material to  the flask by keeping the
                syringe needle just above the surface of the
                acetone.   Caution should be observed to be sure that
              ,  the standard material falls dropwise directly into
                the acetone without contacting the inner wall of the
                volumetric flask.

        7.3.1.3  Reweigh, dilute to volume,  stopper,  then mix by
                inverting  the  flask several times.  Calculate the
                concentration  in milligrams per milliliter from the
                net gain in weight.   Final  concentration should be
                between 0.800  -  1.50 mg/mL.

7.3.2   Larger volumes of standard solution may be prepared at the
       discretion of the analyst.  When compound purity is assayed
       to be 96% or greater, the weight can be used without
       correction to calculate the concentration of the stock
       standard.               -

7.3.3  Commercially prepared stock standards can be used at any
       concentration if they are certified by the manufacturer or by
       an independent source.   When purchasing commercially prepared
       stock standards, every effort should be made to avoid
       solutions  prepared  in methanol  (chloral  hydrate is an
       exception,  Sect. 7.3.3.1).  Methanol can cause degradation of
       most of the haloacetonitriles.   In  addition,  some commercial
       suppliers  have reported instability with solutions of
       simazine and atrazine prepared in methanol (18).   For these
       reasons, acetone should be used as  the primary solvent for
       stock standard and primary dilution  standard  preparation and
                         551.1-13

-------
7.3.4
        all  sources of methanol  introduction into  these  acetone
        solutions should be eliminated.

        7.3.3,1   It is extremely difficult to acquire  chloral  hydrate
                 in its pure form since  it is classified as  a
                 controlled substance.   Consequently,  if pure  chloral
                 hydrate cannot  be acquired,  a commercially  prepared
                 solution of this analyte  (most often  at 1.0 mg/mL)
                 must be purchased.   Most  manufactures provide
                 certified chloral  hydrate solutions in  methanol.
                 Since chloral hydrate is  unstable, standards  from  a
                 separate vendor musft be utilized  to confirm the
                 accuracy of the primary supplier's solution.

        Outside  source stock solutions,  which are  independently
        prepared or purchased from ah outside source different from
        the  source  for the original  stock  standard solutions,  must  be
        used as  a means of verifying the accuracy  of the original
        stock standard solutions for all analytes.  Prepare  a
        dilution of both stocks  in  acetone and perform a final
        dilution in MTBE such that  each  stock dilution is  at the same
        concentration.   Analyze  as  outlined  in Section 11.3.   The
        relative percent difference  (RPD as  defined below) between
        the  analytes'  response (area counts)  from  both solutions
        should not  exceed 25% for any one  analyte.  The  RPD  must be
        less than 20% for 90% or greater of  the total  number of
        target analytes analyzed.
                RPD =
                         (DUP 1 - DUP, 2)
                       ((DUP 1 + DUP 2)  / 2)
                                               X 100
       7.3.4.1
                If this criteria cannot be met, a third outside
                source should be purchased and tested in the same
                manner.  When two sources of stock solutions agree,
                the accuracy of the stock solutions is confirmed.
                This should be done prior to preparing the primary
                dilution standards.

7.3.5  Stock Solution of Surrogate - Prepare a stock solution of the
       surrogate standard in acetone by weighing approx. 0.010 g
       decafluorobiphenyl in a 10-ml. volumetric flask.  When diluted
       to volume this yields a concentration of 1.00 mg/mL. «
       Alternate surrogate analytes,may be selected provided they
       are similar in analytical behavior to the compounds of
       interest, are highly unlikely to be found in any sample, and
       do not coelute with target analytes.

7.3.6  Stock Solution of Internal Standard (IS) - Use of an IS is
       optional when MTBE is the extraction solvent but mandatory if
       pentane is used.  This is due to the high volatility of
       pentane when compared to MTBE (see boiling points,  Sect.

                         551.1-14

-------
             6.9.2.1).   Prepare an internal  standard stock solution of
             bromofluorobenzene (BFB)  in acetone.   Since this compound is
             a  liquid  at room temperature,  the  procedure outlined  in
             Sections  7.3.1.1 through  7.3.1.3  should be  followed but add
             approximately 65 //L of neat BFB rather than 10 fji as
             specified  in  7.3.1.2.   When diluted  to volume this yields a
             concentration near 10.0 mg/mL.  Alternate  internal standard
             analytes may  be  selected  provided  they are  highly unlikely to
             be  found  in any  sample and  do  not  coelute with target
             analytes.

     7.3.7   Transfer the  stock standard solutions  into  Teflon-lined screw
             cap  amber  bottles.   Store at 4°C or less and  protect from
             light.  Stock standard solutions should be  checked frequently
             for  signs  of  degradation or evaporation, especially just
             prior to preparing  calibration  standards from them.

     7.3.8   When stored in a freezer at < -10°C, the THM  stock standards
             have been  shown  to  be  stable for up to  six  months.  The other
             analyte stock standards, with the  exception of chloral
             hydrate, have  been  shown to be  stable  for at  least four
             months when stored  in  a freezer (<-10°C).   Chloral hydrate
             stock standards, when  stored in a  freezer (<-10°C), have  been
             shown to be stable  for  two  months, however, since freezers
             can hold at various temperatures below  -10°C, fresh chloral
             hydrate standards should be initially prepared weekly, until
             the stability of this  analyte is determined for a specific
             laboratory setting.

7.4  PRIMARY DILUTION STANDARDS (PDS)  - Two separate groups of primary
     dilution standards must be prepared; one set in acetone for all the
     method analytes except chloral hydrate and the second set in
     methanol for chloral  hydrate.  Although preparation of separate
     chloral hydrate standards may seem laborious,  due to the stability
     problems encountered with this analyte, making fresh chloral  hydrate
     primary dilution  standards is more efficient.   Prepare primary
     dilution standards by combining and diluting stock standards  in
     acetone (methanol  for chloral hydrate).  The primary dilution
     standards  should  be prepared such that when  25 //L of this primary
     dilution standard are added to 50 ml of buffered/dechlorinated
     reagent water  (Sect 10.1.2),  aqueous concentrations will bracket the
     working concentration range.   Store the primary dilution standard
     solutions  in vials or bottles, with caps  using TFE  faced liners,  in
     a  freezer  (<-10°C) with  minimal headspace  and  check frequently for
     signs  of deterioration or evaporation,  especially just  before
     preparing  calibration standards.   The  same comments on  storage
     stability  in Sect.  7.3.8 apply to  primary  dilution  standards.

     7.4.1   SURROGATE  PRIMARY DILUTION  STANDARD - Dilute 500  //L of the
            surrogate  stock solution to  volume  with acetone  in a 50-mL
            volumetric  flask.   This yields  a primary dilution standard at
            10.0 //g/mL.  Addition  of 50  //L  of this  standard to 50  ml of

                              551.1-15

-------
            aqueous sample yields a final concentration in water of 10.0
            //g/L.  This solution is fortified into the aqueous sample
            prior to extraction of all calibration standards (Sect.
            10.1.3), quality control samples (Sect. 9), laboratory
            reagent blanks (Sect. 9.3.1) and actual field samples (Sect.
            11.1.3) in the extraction set.

     7.4.2  INTERNAL STANDARD (IS) PRIMARY DILUTION STANDARD - Prepare a
            IS primary dilution standard at 100 /vg/mL by diluting the
            appropriate amount of internal standard stock solution (500
            fjl if stock is 10.0 mg/ml) to volume with acetone in a 50-mL
            volumetric flask.  When 10 //L of this solution are added to
            1.0 mL of extract, the resultant final concentration is 1.00
            /yg/mL.  The internal standard is used in order to perform an
            internal standard calibration and is added to an analytically
            precise volume of the extract following extraction.  This
            solution is added to all extracts.

     7.4.3  Reserve approximately a 10 mL aliquot of the same lot of both
            the acetone and methanol used in the preparation of the
            primary dilution standards,  When validating the accuracy of
            the calibration standards (Sect. 7.3.4), fortify a laboratory
            reagent blank with 25 fjl of both the acetone and the methanol
            which was used to prepare the primary dilution standards.
            Analysis of this laboratory reagent blank will confirm no
            target analyte contamination in the solvents used to prepare
            the primary dilution standards.

7.5  LABORATORY PERFORMANCE CHECK SOLUTION (LPC) - To insure proper
     instrument performance, a Laboratory Performance Check Solution is
     prepared.  This solution is prepared in MTBE for direct injection on
     the GC and is used to evaluate the parameters of instrument
     sensitivity, chromatographic performance, column performance and
     analyte breakdown.  These parameters are listed in Table 7 along
     with the method analytes utilized to perform this evaluation, their
     concentration in MTBE and the acceptance criteria.  To prepare this
     solution at the concentrations listed in Table 7, a double dilution
     of the analyte stock solutions must be made.  First prepare a
     primary stock dilution mix at 1000 times the concentrations listed
     in Table 7, by adding the appropriate volume of each stock solution
     to a single 50-mL volumetric flask containing approximately 25 mL of
     MTBE.  Dilute to volume with MTBE,  Then the LPC working solution is
     prepared in MTBE by diluting 50 //L of the primary stock dilution mix
     in MTBE to 50-mL in a volumetric flask.  The best way to accomplish
     this is to add approximately 48 mL MTBE to the 50-mL volumetric
     flask and add 50 jjl of the primary stock dilution mix, then dilute
     to volume with MTBE.  Store this solution in a vial or bottle, with
     TFE faced cap, in a freezer (<-10°C) with minimal headspace and
     check frequently for signs of deterioration or evaporation.

     7.5.1  If a laboratory is not conducting analyses for the high
            molecular weight pesticides and herbicides, a modified LPC

                              551.1-16;

-------
                 may  be  prepared.   This  modified  LPC can omit the endrin
                 analyte breakdown  component  as well as the  resolution
                 requirement  for bromacil  and  alachlor under column
                 performance.   In addition, substitute analytes  in place of
                 nndane for  the sensitivity check and
                 hexachlorocyclopentadiene for chromatographic performance can
                 be selected.   These substitute compounds must meet the same
                 criteria as  listed in table 7 with the concentration for
                 sensitivity  check near  the substitute analyte's EDL and the
                 concentration  for chromatographic performance near 50 times
                 the  substitute analyte's EDL.  The column performance
                 criteria for resolution between bromodichloromethane and
                 trichloroethylene cannot be modified.

          7.5.2  If pentane is  selected  as an alternate extraction solvent the
                 LPC must also be prepared in pentane.

8.   SAMPLE COLLECTION.  PRESERVATION.  AND STORAGE

     8.1  SAMPLE VIAL PREPARATION

          8.1.1  To conduct analyses for the total analyte list,  two sets of
                 60-mL vials must be prepared for sampling.   One  set of vials
                 prepared for the analysis of all  target  analytes except
                 chloral  hydrate,  contains ammonium chloride as  a
                 dechlorinating agent.   Due to concerns over low  recoveries
                 for chloral hydrate in matrices  preserved with  ammonium
                 chloride (Sect. 8.1.2),  a separate sample set must  be
                 collected and preserved  with  sodium  sulfite. Both  sets of
                 vials are prepared  as  follows.

                 8.1.1.1   Using the  homogeneous phosphate        ;
                          buffer/dechlorinating agent mixtures prepared  in
                          Sect.  7.1.7.4,  0.60  g of the  appropriate mixture  are
                          added to the  corresponding vials.

          8.1.2  If the sample assay is for only the THMs  and/or  solvents,
                 either dechlorinating  agent can be added.   However,  sodium
                 sulfite  promotes the decomposition of  the haloacetonitriles,
                 l,l-dichloro-2-propanone,  l,l,l-trichloro-2-propanone   and
                 chloropicrin  and therefore  ammonium chloride must be used  as
                 the dechlorination  reagent  in their analysis.  In addition,
                 some  fortified  matrices, dechlorinated with  ammonium
                 chloride, have  displayed recoveries of chloral hydrate which
                 have  been up  to 50% lower  than expected, when compared to  the
                 same  sample matrix dechlorinated with sodium sulfite.   In
                 other matrices, recoveries have been consistent  regardless of
                 dechlorinating  agent.  The reason for these  differences has
                 not been determined.   Due to this uncertainty, a separate
                 sample,  dechlorinated  with 100 mg/L sodium sulfite must be
                 collected for the analysis of chloral hydrate.


                                  551.1-17

-------
     8.1.3  The dechlorinating agents, if not added within the
            homogeneous mixture of the; buffer, must be added to the
            sampling vials as a dry salt.  Solutions of the
            dechlorinating agents should not be used due to concerns over
            the stability of these salts dissolved in solution and the
            potential chemical interactions of aqueous solutions of these
            salts with the dry phosphate buffer.

     8.1.4  Samples must contain either 100 mg/L ammonium chloride or
            sodium sulfite, as appropriate for the analysis being
            performed.  This amount will eliminate free chlorine residual
            in typical chlorinated drinking water samples.  If high
            chlorine doses are used, such as in a maximum formation
            potential test, additional dechlorinating reagent may be
            required.

8.2  SAMPLE COLLECTION

     8.2.1  Collect all samples in duplicate.  Fill sample bottles to
            just overflowing but ;take care not to flush out the buffer/
            dechlorination reagents.  No air bubbles should pass through
            the sample as the bottle is filled, or be trapped in the
            sample when the bottle is sealed.

     8.2.2  When sampling from a water tap, open the tap and allow the
            system to flush until the water temperature has stabilized
            (usually about 3-5 min).  Remove the aerator and adjust the
            flow so that no air bubbles are visually detected in the
            flowing stream.

     8.2.3  When sampling from an open body of water, fill a 1-quart
            wide-mouth glass bottle or 1-liter beaker with sample from a
            representative area, and carefully fill duplicate 60-mL
            sample vials from the container.

     8.2.4  The samples must be chilled to 4°C on the day of collection
            and maintained at that temperature until analysis.  Field
            samples that will not be received at the laboratory on the
            day of collection must be packaged for shipment with
            sufficient ice to ensure they will be at 4°C on arrival at
            the laboratory.  Synthetic ice (i.e. Blue ice) is not
            recommended.

8.3  SAMPLE STORAGE/HOLDING TIMES     :

     8.3.1  Store samples at 4°C and extracts in a freezer (<-10°C) until
            analysis.  The sample storage area must be free of organic
            solvent vapors.
                              551.1-18

-------
          8.3.2  Extract all samples within 14 days of collection and analyze
                 within 14 days following extractidn. This applies to either
                 MTBE or pentane extracts). Samples not analyzed within these
                 time periods must be discarded and replaced.

9.   QUALITY CONTROL

     9.1  Each laboratory that uses this method is required to operate a
          formal quality control (QC) program.  Minimum QC requirements
          include the laboratory performance check standard,  initial
          demonstration of laboratory capability,  method detection limit
          determination, analysis of laboratory reagent blanks,  continuing
          calibration check standard, laboratory fortified sample matrices,
          field duplicates and monitoring surrogate and/or internal  standard
          peak response in each sample and blank.   Additional quality control
          practices may be added.
                 i
     9.2  ASSESSING.INSTRUMENT SYSTEM - LABORATORY PERFORMANCE CHECK STANDARD
          (LPC).  Prior to any sample analyses,  a  laboratory  performance check
          standard  must be analyzed.   The LPC sample contains compounds
          designed  to indicate appropriate instrument sensitivity, endrin
          breakdown,  column performance (primary column),  and chromatographic
          performance.   LPC sample components and  performance criteria are
          listed in Table 7.   Inability to demonstrate acceptable instrument
          performance indicates the need for revaluation  of  the instrument
          system.   The  sensitivity requirement is  based on the Estimated
          Detection Limits (EDLs).published in this method.   If  laboratory
          EDLs differ from those listed in this  method,  concentrations of the
          LPC standard  must be adjusted to be compatible with the laboratory
          EDLs.   If endrin breakdown  exceeds 20  %,  the problem can most likely
          be solved by  performing routine maintenance on the  injection port
          including replacing the injection port sleeve, and  all  associated
          seals  and septa.   If column or chromatographic performance  criteria
          cannot be met,  new columns  may need to be installed, column flows
          corrected,  or modifications adapted to the oven  temperature program.
          During early  method development work,  significant chromatographic
          and column  performance problems were observed  while using  a DB-1
          column which  had been used  for several years for drinking water
          extract analysis.   By installing a new DB-1  column,  these
          performance problems  were overcome.   If  the  columns to  be used for
          this method have been used  for several years or  have had extended
          use with  extracts  from harsh  sample  matrices (i.e.  wastewater,
          acidified sample extracts,  hazardous waste samples)  it  may  be
          difficult to  meet  the criteria established for this LPC standard and
          column replacement  may be the  best alternative.

     9.3   LABORATORY  REAGENT  BLANKS (LRB).   Before  processing any samples, the
          analyst must  analyze  an  LRB to demonstrate that  all glassware  and
          reagent interferences are under control.   In addition,  each  time a
          set of samples  is extracted or reagents  are  changed, a  LRB  must be
          analyzed.   If the  LRB produces a peak  within the retention  time
          window of any analyte (Sect.  12.2)  preventing  the quantitation  of

                                   551.1-19

-------
     that analyte, determine the source of the contamination and
     eliminate the interference before processing samples.  LRB samples
     must contain the appropriate buffer for the target analytes
     (buffered/NH4Cl  dechlorinated and/or buffered/Na2S03 dechlorinated
     reagent water).

     9.3.1  Prepare the two LRBs in the appropriate buffered/
            dechlorinated reagent water*  Add 50 //L of surrogate primary
            dilution standard (Sect. 7.4.1) to each blank and follow the
            procedure outlined in Sect. 11.2.

9.4  INITIAL DEMONSTRATION OF CAPABILITY (IDC)

     9.4.1  Preparation' of the IDC Laboratory Fortified Blank (LFB).
            Select a concentration for each of the target analyte which
            is approximately 50 times the EDL or close to the expected
            levels observed in field samples.  Concentrations near
            analyte levels in Table 3.A are recommended.  Prepare a LFB
            by adding the appropriate concentration of the primary
            dilution standard (Sect. 7.4) to each of four to seven 50 mL .
            aliquots of buffered/NH4Cl  dechlorinated reagent water.
            Separate Na,S03 preserved matrices  need  not  be analyzed
            (Sect. 9.4.1.1).   Analyze the aliquots according to the
            method beginning in Section 11.

            9.4.1.1  Chloral  hydrate is included in the buffered/NH4Cl
                     dechlorinated reagent water, containing all the
                     other target analytes since no matrix induced
                     recovery problems have been found from reagent water
                     preserved with NH4C1.

     9.4.2  Following procedural  calibration standard analysis and
            subsequent instrument calibration, analyze a set of at least
            seven IDC samples and calculate the mean percent recovery (R)
            and the relative standard deviation of this recovery (RSD).
            The percent recovery is determined as the ratio of the
            measured concentration to the actual fortified concentration.
            For each analyte, the mean recovery value must fall  within
            the range of 80% to 120% and the RSD must not exceed 15 %.
            For those compounds that meet these criteria, performance is
            considered acceptable, and sample analysis may begin.  For
            those compounds that fail  these criteria, this procedure must
            be repeated using eight fresh samples until  satisfactory
            performance has been demonstrated.

     9.4.3  The initial  demonstration of capability is used primarily to
            preclude a laboratory from analyzing and reporting unknown
            samples without obtaining some experience with an unfamiliar
            method.  It is expected that as laboratory personnel gain
            experience with this method,  the quality of data will  improve
            beyond those specified in Sect. 9.4.2.


                              551.1-20 i

-------
      9.4.4   METHOD  DETECTION LIMITS (MDL).   Prior to the analysis of any
             field samples  the method  detection  limits must be determined.
             Initially,  estimate  the concentration of an  analyte which
             would yield a  peak equal  to  5 times the baseline  noise and
             drift.   Prepare  a primary dilution  standard  with  analyte
             concentrations at 1000  times this level  in acetone (or
             methanol  for chloral  hydrate).

             9.4.4.1   Prepare a 500  mL aliquot of buffered/ammonium
                      chloride dechlorinated  reagent  water.  Fill  a
                      minimum of  seven replicate,  60-mL vials  with 50 ml
                      of the  buffered/dechlorinated  (NH4C1) reagent water.


             9.4.4.2   Fortify the  50 ml buffered/dechlorinated (NH4C1)
                      reagent water  with  50 fjl of both the MDL concentrate
                      prepared in  acetone and the chloral  hydrate  MDL
                      concentrate  in methanol. •Separate  preparation  of a
                      reagent water  containing Na2S03 as the
                      dechlorinating agent for chloral  hydrate MDL
                      determination  is  not necessary.  (See Sect. 9.4.1.1)

             9...4.4.3   Extract and  analyze these  samples as outlined  in
                      Section 11.  MDL  determination  can  then  be performed
                      as  discussed in  Sect. 13.1.

9.5  LABORATORY FORTIFIED  BLANK  (LFB).   Since this method utilizes
     procedural calibration  standards, which are  fortified reagent water,
     there is no difference  between the  LFB and  the  continuing
     calibration check  standard.  Consequently,   there  is  not  a
     requirement .for  the analysis of  an  LFB.  However, the criteria
     established for  the continuing calibration  check  standard  (Sect.
     10.4) should be  evaluated as the  LFB.

9.6  LABORATORY FORTIFIED  SAMPLE MATRIX  (L.FM)..   The  laboratory must  add
     known concentrations  of analytes  to a minimum of  10% of  the  routine
     samples or one fortified  sample per sample  set, whichever is
     greater, for both NH4C1  and  Na2S03 dechlorinated  sample matrices.
     The concentrations should be equal  to or greater than the background
     concentrations in the sample selected for fortification.   Over time,
     samples from all routine  sample sources should be fortified.  By
     fortifying sample matrices and calculating  analyte recoveries, any
     matrix induced analyte  bias is evaluated.   When an analyte recovery
     falls outside the acceptance criteria outlined below, a  bias is
     concluded and that analyte for that matrix   is reported to the data
     user as suspect.

     9.6.1  First,  follow the procedure outlined in Sect. 11.1

     9.6.2  Next,  prepare thei LFM by adding  50/yL of an acetone based
            standard solution into the remaining 50 mL of the  buffered/
            NH4C1 dechlorinated sample matrix in the vial  in which it was

                              551.1-21

-------
            sampled.  This sample vial will have had the required amount
            of aqueous sample- removed as specified in Sect. 11.1.2.  Add
            50 /jl of surrogate primary dilution standard (Sect. 7.4.1)
            and follow procedure outlined in Sections 1.1 and 12.

     9.6.3  When chloral hydrate is being determined, prepare the LFM by
            adding 50 jul of a methanol based chloral hydrate standard
            solution into 50 ml_ of the buffered/Na2S03 dechlorinated
            sample matrix in the vial in which it was sampled.  Add 50 /vL
            of surrogate primary dilution standard (Sect. 7.4.1) and
            follow procedure outlined in Sections.11 and 12.

     9.6.4  Calculate the percent recovery, R, of the concentration for
            each analyte, after correcting the total measured
            concentration, A, from the fortified sample for the
            background concentration, B, measured in the unfortified
            sample, i.e.:
                             R = 100 (A - B) / C,

            where C is the fortifying concentration.  The recoveries of
            all analytes being determined must fall between 75 % and 125
            % and the recoveries of at least 90% of these analytes must
            fall between 80 % and 120 %.  This criteria is applicable to
            both external and internal standard calibrated quantitation.

     9.6.5  If a recovery falls outside of this acceptable range, a
            matrix induced bias can be assumed for the respective analyte
            and the data for that analyte in that sample matrix must be
            reported to the data user as suspect.

     9.6.6  If the unfortified matrix has analyte concentrations equal to
            or greater than the concentration fortified, a duplicate
            sample vial needs to be fortified at a higher concentration.
            If no such sample is available the recovery data for the LFM
            sample should not be reported for this analyte to the data
            user.

9.7  FIELD DUPLICATES (FD1 and FD2).  The laboratory must analyze a field
     sample duplicate for a minimum of 10% of the total number of field
     samples or at least one field sample duplicate per sample set,
     whichever is greater.  Duplicate results must not reflect a relative
     percent difference (RPD as defined below) greater than 25% for any
     one analyte and the RPD for 90% of the analytes being determined
     must be less than 20%.
     RPD
              (FD1 - FD2)
            ((FD1 + FD2) / 2)
X 100
                              551.1-22

-------
     where FD1 and FD2 represent the quantified concentration on an
     individual analyte for the initial and duplicate field sample
     analysis, respectively.  If this criteria is not met the analysis
     must be repeated.  Upon repeated failure, the sampling must be
     repeated or the analyte out of control must be reported as suspect
     to the data user.

9.8  ASSESSING SURROGATE RECOVERY

     9.8.1  The surrogate analyte is fortified into the aqueous portion
            of all  calibration standards, quality control samples and
            field samples.  By monitoring the surrogate response, the
            analyst generates useful quality control information from
            extraction precision through quantitative analysis.
            Deviations in surrogate recovery may indicate an extraction
            problem.   If using external standard calibration the
            surrogate retention time functions as a reference for
            identification of target analytes.

     9.8.2  Using the mean surrogate response from the calibration
            standard  analyses (Cal<.R), determine the surrogate percent
            recovery  (%RECS)  in  all  calibration  standards,  LFBs,  and
            LFMs,  and field samples.  This recovery is calculated by
            dividing  the surrogate response from the sample (SamSR)  by
            the mean  response from the initial calibration standards
            (Sect.  10.2 or 10.3) and multiplying by 100,  as shown below.
                                                        *,
                                  SamSR
                     % RECS  =	x  100
                                  Ca1SR

            Recoveries must  fall  within the range  of 80%  to 120%.   If a
            sample  provides  a recovery outside of  this range,  the extract
            must  be reanalyzed.   If upon reanalysis,  the  recovery
            continues to fall  outside the acceptable range a fresh sample
            should  be extracted  and analyzed.   If  this is not  possible
            the data  for all  the analytes from this  sample should be
            reported  to the  data user as suspect due to  surrogate
            recovery  outside acceptable limits.

     9.8.3   If consecutive samples  fail  the  surrogate  response  acceptance
            criterion,  immediately  analyze a continuing calibration
            standard.

            9.8.3.1   If the  continuing  calibration  standard  provides a
                     recovery  within the acceptable  range of 80%  to  120%,
                     then  follow procedures  itemized  in Sect. 9.8.2  for
                     each  sample  failing the surrogate response
                     criterion.

            9.8.3.2   If the  check standard provides  a  surrogate recovery
                     which falls  outside the acceptable range or  fails

                              551.1-23

-------
                     the acceptance criteria specified in Sect. 10.4 for
                     the target analytes, then the analyst must
                     recalibrate, as specified in Sect. 10.

9.9  ASSESSING THE INTERNAL STANDARD (IS)

     9.9.1  When using the internal standard calibration procedure, the
            analyst must monitor the internal standard response (peak
            area or peak height) of all samples during each analysis day.
            The internal standard response should not deviate from mean
            internal standard response of the past five continuing '
            calibration standards by more than 20%.             ..

     9.9.2  If > 20% deviation occurs wiith an individual extract,
            optimize instrument performance and inject a second aliquot
            of that extract.

            9.9.2.1  If the reinjected aliquot produces an acceptable
                     internal standard response, report results for that
                     aliquot.

            9.9.2.2  If a deviation of greater than 20% is obtained for
                     the reinjected extract, analysis of a calibration
                     check standard must be performed (Sect. 10.4).

     9.9.3  If consecutive samples fail this IS response acceptance
            criterion, immediately analyze a calibration check standard.

            9.9.3.1  If the check standard provides ,a response factor
                     (RF) within 20% of the predicted value for the
                     internal standard and the criteria for all the
                     target analytes as specified in Sect. .10.4 is met,
                     the previous sample(s) failing the IS response
                     criteria need to be reextracted provided the sample
                     is still available.1  In the event that reextraction
                     is not possible, report results obtained from the
                     reinjected extract (Sect 9.9.2) but annotate as
                     suspect due to internal standard recovery being
                     outside acceptable limits.

            9.9.3.2  If the check standard provides a response factor
                     which deviates more than 20% of the predicted value
                     for the internal standard or the criteria for the
                     target analytes, as specified in Sect 10.4 are not
                     met, then the analyst must recalibrate, as specified
                     in Sect. 10.3 and a;ll samples analyzed since the
                     previous calibration check standard need to be
                     reanalyzed.

9.10 CONFIRMATION COLUMN ANALYSIS.  If a positive result is observed on
     the primary column, a confirmation analysis should be performed
     using either the confirmation column or by GC/MS.

                              551.1-24

-------
     9.11 The laboratory may adapt  additional  quality  control  practices  for
          use with this method.  The specific  practices  that are most
          productive depend upon.the needs of  the laboratory and the nature of
                          the samples.  For example, field reagent blanks may
                          be used to assess contamination of samples under
                          site conditions, transportation and  storage.

     9.12 Quality control samples (QCS) from an outside  source, as defined in
          i>ect.  3.12, should be analyzed at least quarterly.

10.  CALIBRATION AND STANDARDIZATION

     10.1 PREPARATION OF CALIBRATION STANDARDS

          10.1,1 Five calibration standards are required.  One should contain
                 the analytes  at a concentration near to but greater than the
                 method detection limit (Table 2}  for each compound;  the
                 others should  be evenly distributed throughout the
                 concentration  range expected  in samples or define  the working
                 range  of the detector.  Guidance  on the number of  standards
                 is  as  follows:   A minimum of  three  calibration standards are  '
                 required  to calibrate  a range of  a  factor of 20 in
                 concentration.   For a  factor  of 50  use at least four
                 standards, and  for  a  factor of 100  at  least  five standards
                 For  example, if the MDL is 0.1 /zg/L, and a  sample
                 concentrations  are  expected to range from 1.0  jug/I to  10 0
                jjg/L,  aqueous standards should be prepared at  0 20 uq/L  *0 80
                 /ig/L,  2.0 ./ig/L,  5.0 /tg/L,  and  15.0  ftg/l.-            *'  '

         10.1.2  As a means of eliminating any  matrix effects due to  the  use
                 of the phosphate buffer and dechlorinating agents  the
                 procedural calibration  standards are prepared  in reagent
                water which has been buffered  to pH 5.2  and dechlorinated
                with ammonium chloride.  To prepare this
                buffered/dechlorinated  reagent water,  add 5.0  g of phosphate
                buffer/dechlorinating  agent (Sect 7.1.7.4, ammonium chloride
                type) to 500 mL of  reagent water (Sect.  7.2).

         10.1.3  Next  add 25 /yL of the desired concentration primary dilution
                standards (acetone and methanol based,  Sect.  7.4)  to a 50 ml
                aliquot of the  buffered/dechlorinated reagent water in a 60-
                 r !u  \  ,   a 50~/zL micro syringe and rapidly inject 25 ul
                of the  standard into the middle point of the water volume
                M  2Ve  ^erneed1e as quickly  as possible after injection.'
                Next,  add 50 //L  of the  surrogate standard solution  (Sect
                7.4.1)  in the same manner.  Mix by slowly and carefully
                inverting the sample vial  two  times  with minimal sample
                agitation.  Aqueous  standards  must be prepared  fresh  daily
                and  extracted immediately  after preparation  (Section  11.2).

                10.1.3.1  By including chloral  hydrate into the  total  NH,C1
                         analyte  matrix, .a  separate  calibration standard

                                  551.1-25

-------
                     analysis for Na2S03 preserved reagent water
                     fortified with chloral  hydrate  is avoided.   Chloral
                     hydrate is included in'the buffered/NH4Cl
                     dechlorinated reagent water,  containing all  the
                     other target analytes since no  matrix induced
                     recovery problems have been found from reagent water
                     preserved with NH4C1.  Warning!  Do not attempt to
                     analyze  chloral  hydrate in field samples preserved
                     with NH4C1,  low recoveries may  result due to matrix
                     effects.

     10.1.4 CAUTION - DO NOT prepare procedural calibration standards  in
         1   a volumetric flask and transfer the sample to an extraction
            vial either directly for weight determination of volume or
            into a graduated cylinder with a subsequent additional
            transfer into the extraction vial.  Volatility experiments
            reflected as much as,a 30 % loss in volatile low molecular
            weight analytes following such transfers.  All fortified
            samples and field samples mu$t be extracted in the vial or
            bottle in which they were fortified and collected.

10.2 EXTERNAL STANDARD CALIBRATION PROCEDURE

     10.2.1 Extract and analyze each calibration standard according to
            Section 11 and tabulate peak height or area response versus
            the concentration of the standard.  The results are used to
            prepare a calibration curve for each compound by plotting the
            peak height or area response versus the concentration.  This
            curve can be defined as either first or second order.
            Alternatively, if the ratio of response to concentration
            (response factor) is constant over the working range  (< 10%
            relative standard'deviation,[RSD]), linearity through the
            origin can be assumed, and the average ratio or calibration
            factor can be used in place of a calibration curve.

     10.2.2 Surrogate analyte recoveries must  be verified as detailed in
            Sections 9.8.

10.3 INTERNAL STANDARD (IS) CALIBRATION PROCEDURE

     10.3.1 Extract each calibration standard  according to Section  11.
            Remove a 1.00 mL  portion of ;the MTBE or pentane extract from
            the sample extraction vial and place this  into a  2.0-mL
            autosampler vial.  To this extract, add the 10 //L of  the
            internal standard primary dilution  standard,  cap  the  vial and
            analyze.  Following  analysis, tabulate  peak height or  area
            responses against concentration for each  compound and  the
            internal standard.   Calculate relative  response factor  (RRF)
            for each compound using Equation  1.
                               551.1-26

-------
Equation 1

         RRF = j


where
                                    (A,.)  (C.)
                 As  = Response for the analyte to be measured
                 Ais = Response for the internal  standard
                 C-s = Concentration of the internal  standard
                 Cs  = Concentration of the analyte to be measured (fJ.g/1)

                 If RF value over the working range is constant (< 10% RSD),
                 the  average RF can be used for calculations.  Alternatively,
                 the  results can be used to plot a calibration curve of
                 response versus analyte ratios, As/Ais vs. Cs.

     10.4 CONTINUING  CALIBRATION CHECK STANDARD

          10.4.1 Preceding each analysis set, after every tenth sample
                 analysis and after the final sample analysis, a calibration
                 standard should be analyzed as a continuing calibration
                 check. These check standards should be at two different
                 concentration levels to verify the calibration curve.  This
                 criteria is applicable to both external  and internal standard
                 calibrated quantitation.  Surrogate and internal  standard
                 recoveries must be verified as detailed in Sections 9.8 and
                 9.9, respectively.

          10.4.2 In order for the calibration to be considered valid, analyte
                 recoveries for the continuing calibration check standard must
                 fall between 75 % and 125 % for all  the target analytes.  The
                 recoveries of at least 90% of the analytes determined must
                 fall between 80% and 120%

          10.4.3 If this criteria cannot be met, the continuing calibration
                 check standard is reanalyzed in order to determine if the
                 response deviations observed from the initial analysis are
                 repeated.  If this criteria still cannot be met then the
                 instrument is considered out of calibration for those
                 specific analytes beyond the acceptance range.  The
                 instrument needs to be recalibrated and the previous samples
                 reanalyzed or those analytes out of acceptable range should
                 be reported as suspect to the data user for all the
                 previously analyzed samples.
11.   PROCEDURE
     11.1 SAMPLE PREPARATION

          11.1.1 Remove samples from storage and allow them to equilibrate to
                 room temperature.

                                   551.1-27

-------
     11.1.2 Remove the vial caps.  Remove a 10 ml volume of the sample.
            Check the pH of this 10 ml aliquot to verify that it is
            within a pH range of 4.5 anc'l 5.5.  If the pH is out of this
            range a new sample must be collected.  Replace the vial caps
            and weigh the containers with contents to the nearest 0.1 g
            and record these weights for subsequent sample volume
            determination.  (See Sect. 11.2.4  for continuation of
            weighing and calculation of true volume).  Alternatively, the
            sample vials may be precalibrated by weighing in 50 ml of
            water and scoring the meniscus on the bottle.  This will
            eliminate the gravimetric step above and in Sect. 11.2.4.

     11.1.3 Inject 50//L of the surrogate analyte fortification solution
            (Sect. 7.4.1) into the sample.  The aqueous concentration of
            surrogate analyte must be tHe same as that used in preparing
            calibration standards (Sect. 9.1.3).  Mix by slowly and
            carefully inverting the sample vial two times with minimal
            sample agitation.

11.2 SAMPLE EXTRACTION                  i

     11.2.1 WITH MTBE AS EXTRACTION SOLVENT

            11.2.1.1 After addition of the surrogate (Sect 11.1.3) add
                     exactly 3.0 mL of MTBE with a type A, TD, transfer
                     or automatic dispensing pipet.

            11.2.1.2 Add 10 g NaCl  or 20 g Na2S04 to the  sample vial.
                     (See Section 13.7 for an important notice concerning
                     the use of NaCl when analyzing for DBFs.)  Recap and
                     extract the NaCl or Na2SO,  /MTBE/sample mixture  by
                     vigorously and consistently shaking the vial by hand
                     for 4 min.  Invert the vial and allow the water and
                     MTBE phases to separate (approx. 2 min).

                     If a series of samples are being prepared for
                     extraction using Na2S04, immediately  after the
                     addition of the Na2504, the sample should be
                     recapped, agitated and placed in a secure horizontal
                     position with the undissolved Na,SO,  distributed
                     along the length of the vial.   If trie vial  is left
                     in a vertical  position, while additional samples
                     have solvent and salt added,  the Na2S04 will
                     solidify in the bottom of the vial  and it will not
                     dissolve during sample extraction.

                     NOTE:  Previous versions of this method call for the
                     addition of the salt by "shaking the vial
                     vigorously" before the MTBE has been added.   Please
                     make a note that this procedural  order has been
                     changed in an effort to minimize volatile analyte
                     losses.

                              551.1-28

-------
       11.2.1.3 By using a disposable Pasteur pipet  (Sect. 6.2),
                transfer a portion of the solvent phase from the 60-
                mL vial to an autosampler vial (Sect. 6.2).  Be
                certain no water has carried over onto the bottom of
                the autosampler vial.  If a dual phase appears in
                the autosampler vial, the bottom layer can be easily
                removed and discarded by using a Pasteur pipet.  The
                remaining MTBE phase may be transferred to a second
                autosampler vial as a backup extract or for separate
                confirmation analysis.  Approximately 2.5 ml of the
                solvent phase can be conveniently transferred from
                the original 3 ml volume.

                11.2.1.3.1  If using  an  internal  standard
                            quantitation,  the  extract transfer into
                            the  autosampler  vial  must be performed
                            in a quantitative  manner.  This may be
                            done using a  1.00  ml  syringe or a  2.00-
                            mL graduated  disposable  pipet  to
                            accurately transfer  1.00 mL o.f sample
                            extract to the autosampler vial  where  10
                           //L of internal standard  primary dilution
                            standard  (Sect.  7.4.2) solution can be
                            added.

11.2.2 WITH PENTANE AS EXTRACTION SOLVENT

       11.2.2.1 After addition of the surrogate (Sect 11.1.3)  add
                exactly 5.0 mL of pentane with a type A,  TD,
                transfer or automatic dispensing pipet.

       11.2.2.2 Add 20 g Na2S04  to the sample vial.   Recap  and
                extract the Na2S04/pentane/sample mixture by
                vigorously and consistently  shaking  the vial by hand
                for 4 min.   Invert the vial  and allow the  water and
                pentane phases to separate (approx.  2 min).  NOTE:.
                Previous versions of  this method call for  the
                addition of NaCl by "shaking the vial vigorously"
                before the pentane has been  added.   Please make a'
                note that this procedural  order has  been  changed in
                an effort to minimize.volatile analyte losses.  If a
                series of samples are being  prepared for  extraction,
                immediately after the addition of the Na2S04, the
                sample should be recapped,  agitated  and placed in  a
                secure horizontal  position with  the  undissolved
                Na2S04 distributed along the length of the vial.  If
                the vial  is left in a vertical  position, while
                additional  samples have  solvent  and  salt  added, the
                Na2SO^ will solidify in the bottom of the vial  and
                it will  not dissolve  during  sample extraction.
                         551.1-29

-------
            11.2.2.3 Using a disposable Pasteur pipet,  transfer a portion
                     of the solvent phase from the 60-mL vial  to an
                     autosampler vial,.  Be certain no water has carried
                     over onto the bottom of the autosampler vial.  If a
                     dual phase appeals in the autosampler vial, the
                     bottom layer can be easily removed and discarded
                     using a Pasteur pipet.  The remaining pentane phase
                     may be transferred to a second autosampler vial as a
                     backup extract or for separate confirmation
                     analysis.

                     11.2.2.3.1 The extract transfer into the
                                autosampler vial must be performed  in a
                                quantitative manner.  This may  be done
                                using a 1.00-mL syringe or a 2.00-mL
                                graduated disposable pipet to accurately
                                transfer 1.00 ml of sample, extract  to
                                the autosampler vial where 10 /A. of
                                internal standard primary dilution
                                standard (Sect. 7.4.2) solution can be
                                added.

     11.2.3 Discard the remaining contents of the sample vial.  Shake off
            the last few drops with short, brisk wrist movements.

     11.2.4 Reweigh the empty vial with the original cap and calculate
            the net weight of sample iby difference to the nearest 0.1 g
            (Sect. 11.1.2 minus Sect. 11.2.4).  This net weight (in
            grams) is equivalent to the volume of water (in ml)
            extracted, Vs.

     11.2.5 The sample extract may be stored in a freezer (<-10°C)  for a
            maximum of fourteen days before chromatographic analysis but
            no more than 24 hours at room temperature (i.e. on  an
            autosampler rack).  Due to the volatility of the extraction
            solvent, if the septum on a vial has been pierced,  the  crimp
            top or screw cap septum needs to be replaced immediately or
            the extract cannot be reanalyzed at a later time.

11.3 SAMPLE ANALYSIS

     11.3.1 The recommended GC operating conditions are described in
            6.9.2.1 and 6.9.2.2 along with recommended primary  and
            confirmation columns.  Retention data for the primary and
            confirmation columns are given in Table 1.

     11.3.2 Inject 2 /zL of the sample extract and record the resulting
            peak response.  For optimum performance and precision,  an
            autosampler for sample injection and a data system  for  signal
            processing are strongly recommended.
                              551.1-30

-------
12.  DATA ANALYSIS AND CALCULATIONS

     12.1 Identify sample components by comparison of retention times to
          retention data from the calibration standard analysis.  If the
          retention time of an unknown compound corresponds, within limits
          (Sect. 12.2), to the retention time of a standard compound, then
          identification is considered positive.

     12.2 The width of the retention time window used to make identifications
          should be based upon measurements of actual retention time
          variations of standards over the course of a day.  Three times the
          standard deviation of a retention time can be used to calculate a
          suggested window size for a compound.  However, the experience of
          the analyst should weigh heavily in the interpretation of
          chromatograms.  Use the initial demonstration of capability
          retention time data as an initial means of determining acceptable
          retention time windows.  Throughout the development of this method a
          retention time window of 1.0 % of the total analyte retention time
          was used.

     12.3 Identification requires expert judgment when sample components are
          not resolved chromatographically, that is, when GC peaks obviously
          represent more than one sample component (i.e., broadened peak with
          shoulder(s) or valley between two or more maxima).  Whenever doubt
          exists over the identification of a peak in a chromatogram,
          confirmation is suggested by the use of a dissimilar column or by
          GC-MS when sufficient concentrations of analytes are present.

     12.4 If the peak response exceeds the linear range of the calibration
          curve, the final  extract should be diluted with the appropriate
          extraction solvent and reanalyzed.  The analyst is not permitted to
          extrapolate beyond the concentration range of the calibration curve.

     12.5 Calculate the uncorrected concentrations (Cf)  of each  analyte in  the
          sample from the response factors or calibration curves generated in
          Sect.  10.2.1 or 10.3.1.  do not use the daily calibration check
          standard to calculate amounts of method analytes in samples.

     12.6 Calculate the corrected sample concentration as:

                Concentration,  /ig/L = C-  x  50  ,
                                            Vs

          where  the sample  volume,  Vs  in  mL,  is  equivalent to  the  net sample
          weight in grams determined in Sect.  11.1.2 and Sect. 11.2.4.

13.   METHOD  PERFORMANCE

     13.1 In a single laboratory,  analyte recoveries from reagent  water with
          MTBE as the extracting solvent,  were determined at three
          concentration levels,  Tables 2A through 4B.  Results from the lowest
          fortified level were  used to determine the analyte MDLs-(11)  listed

                                   551.1-31

-------
     in Table 2.  These MDLs along with the estimated detection limit
     (EDL) were determined in the following manner.   EDLs are provided
     for informational purposes.

     13.1.1 For each analyte, calculate the mean concentration and the
            standard deviation of this mean between  the seven replicates.
            Multiply the student's t-value at 99% confidence and n-1
            degrees of freedom (3.143 for seven replicates)  by this
            standard deviation to yield a statistical estimate of the
            detection limit.  This estimate is the MDL.

     13.1.2 Since the statistical estimate is based  on the precision of
            the analysis, an additional estimate of detection can be
            determined based upon themoise and drift of the baseline as
            well as precision.  This estimate, known as the "EDL" is
            defined as either the MDL or a level of compound in a sample
            yielding a peak in the final extract with a,signal to noise
            (S/N) ratio of approximately 5, whichever is greater.

     13.1.3 These MDL determinations were conducted  on both the primary
            (DB-1) and the confirmation (Rtx-1301) columns and are
            presented in Tables 2.A. through 2.D.

13.2 Analyte recoveries were also determined for reagent water with
     pentane as the extracting solvent.  Two concentration levels were
     studied and the results are presented in Tables 8 and 9.  Results
     from the lowest fortified level were used to determine the analyte
     MDLs (11) listed in Table 8.  These MDLs along  with the estimated
     detection limit (EDL) were determined in a manner analogous to that
     described in Sect. 13.1.1 through 13.1.2.

13.3 In a single laboratory, method precision and accuracy were evaluated
     using analyte recoveries from replicate buffered/dechlorinated (both
     NH,C1  and Na2S03) matrices with MTBE as the extracting solvent.  The
     matrices studied included; fulvic acid fortified reagent water and
     ground water displaying a high CaCO, content.  The results for these
     are presented in Tables 3.A. through 6.B.  These matrices were
     fortified using outside source analyte solutions (except for the
     pesticides and herbicides) to assess accuracy and eight replicate
     analyses were conducted to assess precision.

13.4 Holding time studies were conducted for buffered/dechlorinated
     reagent water and tap water.  Holding studies were also conducted on
     MTBE sample extracts from these two matrices.  Results indicated
     that analytes were stable in these water matrices stored at 4°C.

13.5 MTBE and pentane extracts holding studies indicated the analytes
     were stable for 14 days when stored in a freezer at <-10°C.

13.6 Chromatograms of a fortified, buffered/NH4Cl  dechlorinated reagent
     water extract are presented as Figures 1 through 3.  In the
     chromatograms of Figures 1 and 2, the elution of the method analytes

                              551.1-32

-------
           from a MTBE extract can be seen on the primary DB-1 column and the
           confirmation Rtx-1301 column,  respectively.   Figure 3 shows the
           elution of the method analytes from a pentane extract,  using a
           modified temperature program,  on the primary DB-1 column.  Analyte
           numerical  peak identification, retention time and fortified
           concentrations are presented for information purposes only in Tables
           10,  11 and 12 for Figures 1, 2 and 3,  respectively.

      13.7  IMPORTANT  NOTICE:  All  demonstration data presented in Section'17
           using  MTBE as the extracting solvent,  was obtained using NaCl  as  the
           salt.   A recent report (19)  indicated  elevated recoveries  (via
           synthesis)  of some brominated  DBPs when  NaCl  was  used in the
           extraction process,  due to the inevitable presence of bromide
           impurities in the NaCl.   This  phenomenon has  been confirmed by the
           authors of this method in samples  from chlorinated water systems
           that were  not extracted immediately after the NaCl  was  added.
           Significant effects  can be seen  if extraction is  delayed for as
           little as  15  minutes  after the addition  of the NaCl.   For  this
           reason,  the use of Na?SO, is strongly recommended  over NaCl for MTBE
           extraction  of DBPs.  Although  less  method validation  data  have been
           obtained for  the  Na2S04 option, sufficient data have been  collected
           to indicate that  it  is equivalent  or superior to  NaCl  in salting  out
           the method  analytes, and  has no  observed  negative  effect on
           precision  or  accuracy.

14.  POLLUTION PREVENTION

     14.1  This method  is  a micro-extraction procedure which  uses  a minimal
           amount  of extraction solvent per sample.  This microextraction
           procedure reduces  the  hazards  involved with handling  large volumes
           of potentially  harmful organic solvents needed for conventional
           liquid-liquid extractions.

     14.2  For information about  pollution prevention that may be applicable to
           laboratory operations, consult "Less is Better:  Laboratory Chemical
          Management for  Waste Reduction", available from the American
          Chemical Society's Department of Government Relations and  Science
          Policy, 1155  16th Street N.W.,  Washington, D.C. 20036.

15.  WASTE MANAGEMENT        ,

     15.1 Due to  the nature of this method, there is little  need for waste
          management.  No large volumes of solvents or  hazardous chemicals are
          used.   The matrices of concern are finished drinking water or source
          water,.   However, the Agency requires that laboratory waste
          management  practices be conducted consistent  with  all  applicable
          rules  and regulations,  and that laboratories  protect the air, water,
          and land by minimizing and controlling all releases from fume hoods'
          and bench operations.   Also,  compliance is required with any sewage'
         discharge permits and regulations,  particularly the hazardous waste
          identification rules  and land disposal  restrictions.   For  further
          information on waste  management,  consult  "The Waste Management

                                   551.1-33

-------
          Manual for Laboratory Personnel,",also available from the American
          Chemical Society at the address in Sect. 14.2.

16.  REFERENCES                             '

     1.   Glaze, W.W., Lin, C.C., "Optimization of Liquid-Liquid Extraction
          Methods for Analysis of Organics  in Water", EPA-600/S4-83-052, U.S.
          Environmental Protection Agency, January 1984.

     2.   Richard, J.J., Junk, G.A., "Liquid Extraction for Rapid
          Determination of Halomethanes in Water," Journal AWWA. 69, 62, 1977.

     3.   Reding, R., P.S. Fair, C.J. Shipp, and H.J. Brass, "Measurement of
          Dihaloacetonitriles and Chloropicrin in Drinking Water",
          " Disinfection Byproducts: Current Perspectives ", AWWA, Denver,CO
          1989.

     4.   Hodgeson, J.W., Cohen, A.L. and Collins, J. P., "Analytical Methods
          for Measuring Organic Chlorination Byproducts" Proceedings Water
          Quality Technology Conference (WQTC-16), St. Louis, MO, Nov. 13-17,
          1988, American Water Works Association, Denver, CO, pp. 981-1001.

     5.   Henderson, J.E., Peyton, G.R. and Glaze, W.H. (1976).  In
          "Identification and Analysis of Organic Pollutants in Water" (L.H.
          Keith ed.), pp 105-111.  Ann Arbor Sci. Pub!., Ann Arbor, Michigan.

     6.   Fair, P.S., Barth, R.C., Flesch, J.J. and Brass, H., "Measurement of
          Disinfection Byproducts in Chlorinated Drinking Water," Proceedings
          Water Quality Technology Conference (WQTC 15), Baltimore, MD, None.
          15-20, 1987, American Water Works Association, Denver, CO, pp 339-
          353

     7.   Trehy, M.L. and Bieber, T.I. (1981). In " Advances in the
          Identification and Analysis of Organic Pollutants in Water II" (L.H.
          Keith, ed.) pp 941-975. Ann Arbor Sci. Publ., Ann Arbor, Michigan.

     8.   Oliver, B.G., "Dihaloacetonitriles in Drinking Water: Algae and
          Fulvic Acid as Precursors," Environ. Sci. Techno!, 17, 80, 1983.

     9.   Krasner, S.W., Sclimenti, M.J.  and Hwang, C.J., "Experience with
          Implementing a Laboratory Program to Sample and Analyze for
          Disinfection By-products in a National Study," Disinfection By-
          products: Current Perspectives. AWWA, Denver, CO, 1989.

     10.  Munch, J. W., "Method 525.2-Deterrni nation of Organic Compounds in
          Drinking Water by Liquid-Solid Extraction and Capillary Column
          Chromatography/ Mass Spectrometry" in Methods for the Determination
          of Organic Compounds in Drinking Water; Supplement 3  (1995).
          USEPA, National Exposure Research Laboratory, Cincinnati, Ohio
          45268.
                                   551.1-34

-------
11.  Munch, J.W.,  "Method 524.2- Measurement of Purgeable Organic
     Compounds in Water by Capillary Column Gas Chromatography/ Mass
     Spectrometry" in Methods for the Determination of Organic Compounds
     in Drinking Water; Supplement 3  (1995).  USEPA, National Exposure
     Research Laboratory, Cincinnati, Ohio 45268.

12.  Glaser, J.A., Foerst, D.L., McKee, G.D., Quave, S.A. and Budde, W.L.
     "Trace Analysis for Wastewaters", Environ. Sci. Technol.. 15. 14£6,
     1981.                                                ~~  ~"

13.  ASTM Annual Book of Standards, Part 11, Volume 11.02, D3694-82,
     "Standard Practice for Preparation of Sample Containers and for
     Preservation," American Society for Testing and Materials,
     Philadelphia, PA, 1986.

14.  Bellar, T.A., Stemmer, P., Lichtenburg, J.J.,  "Evaluation of
     Capillary Systems for the Analysis of Environmental Extracts," EPA-
     600/S4-84-004, March 1984.

15.  "Carcinogens-Working with Carcinogens", Publication No. 77-206,
     Department of Health, Education, and Welfare,  Public Health Service,
     Center for Disease Control, National Institute of Occupational
     Safety and Health, Atlanta, Georgia, August 1977.

16.  "OSHA Safety and Health Standards, General Industry", (29CFR1910),
     OSHA 2206, Occupational Safety and Health Administration,
     Washington, D.C. Revised January 1976.

17.  "Safety in Academic Chemistry Laboratories", 3rd Edition, American
     Chemical  Society Publication, Committee on Chemical Safety,
     Washington, D.C., 1979.

18.  Cole, S.,  Henderson, D.  "Atrazine and  Simazine - Product Redesign
     improves  Stability".  The Reporter,  Volume 13, No.  6, 1994,  pg 12.
     Trade publication from Supelco,  Inc.

19.  Xie,  Yuefeng, "Effects of Sodium Chloride on DBP Analytical
     Results,"  Extended Abstract,  Division of Environmental  Chemistry,
     American  Chemical Society Annual Conference, Chicago, IL, Aug. 21-
     26,  1995.
                              551.1-35

-------
TABLE 1.  RETENTION TIME DATA USING MTBE
ANALYTE
Chloroform
1 , 1 , 1-Tri chl oroethane
Carbon Tetrachloride
Trichl oroacetoni tri le
Di chl oroacetoni tri 1 e
Bromodichloromethane
Trichloroethylene
Chloral Hydrate
1 , 1-Dichl oro-2-Propanone
1,1,2-Tri chl oroethane
Chloropicrin
Di bromochl oromethane
Bromochl oroacetoni tri 1 e
1,2-Dibromoethane (EDB)
Tetrachl oroethyl ene
1,1, 1-Tri chl oropropanone
Bromoform
Di bromoacetoni tril e
1 , 2,3-Trichl oropropane
l,2-Dibromo-3-chloropropane (DBCP)
Hexachl orocycl opentadi ene
Trifluralin
Simazine
Atrazine
Hexachl orobenzene
Lindane (gamma-BHC)
Metribuzin
Bromacil
Column Aa
Retention Time
minutes
7.04
8.64
9.94
10.39
12.01
12.42
12.61
13.41
14.96
19.91
23 . 10
23.69
24.03
24.56
26.24
27.55
29.17
29.42
30.40
35.28
40.33
45.17
46.27
46.55 .
47.39
47.95.
150.25
152.09
Column Bb
Retention Time
minutes
7.73
7.99
8.36
10.35
25.21
15.28
11.96
NR c
20.50
25.01
23.69
26.32
29.86
26.46
24.77
28.47
30.36
32.77
31.73
36.11
39.53
45.43
48.56d
48.56d
46.47
49.68
53.92
59.60
                551.1-36

-------
              TABLE  1.   RETENTION TIME DATA USING MTBE (cont'd)
 ANALYTE
                                          Column Aa
                                       Retention Time
                                           minutes
   Column Bb
Retention Time
    minutes
Alachlor
Cyanazine
Heptachlor
Metolachlor
Heptachlor Epoxide
Endrin
Endrin Aldehyde
Endrin Ketone
Methoxychlor
Surrogate:
Decaf luorobiphenyl
Internal Standard:
52.25
53.43
53.72
55.44
58.42
64.15
65.46
72.33
73.53
36.35
31.00
54.38
59.89
53.15
57.07
59.05
65.24
71.56
81.28
76.73
36.28
31.30
 Bromof1uorobenzene
(a)   Column A -     0.25 mm ID x 30 m fused silica capillary with chemically
                    bonded methyl polysiloxane phase (J&W, DB-1, 1.0 /zm film
                    thickness or equivalent).  The linear velocity of the
                    helium carrier is established at 25 cm/sec at 35°C.
               The column oven is temperature programmed as follows:
               [1]  HOLD at 35°C for 22 min
               [2]  INCREASE to 145°C at 10°C/min and hold  at  145°C for 2 min
               [3]  INCREASE to 225°C at 20°C/min and hold  at  225°C for 15 min
               [4]  INCREASE to 260°C at 10°C/min and hold  at  260°C for 30
                    min.. or until all expected compounds have eluted.
               Injector temperature:  200°C
               Detector temperature:  290°C
(b)   Column B -
                    0.25 mm ID x 30 m with chemically bonded 6 %
                    cyanopropylphenyl/94 % dimethyl polysiloxane phase
                    (Restek, Rtx-1301, 1.0 jum film thickness or equivalent).
                    The linear velocity of the helium carrier gas is
                    established at 25 cm/sec at 35°C.
               The column oven is temperature programmed exactly as indicated
               for column A, above.  The same temperature program is utilized
               to allow for simultaneous confirmation analysis.
(c)
     There is no retention time for this analyte since it does not separate
     into a discreet peak on the Rtx-1301.
(d)   Atrazine and simazine coelute on the confirmation column.
                                   551.1-37

-------
   TABLE 2. A.
NH,C1  PRESERVED
METHOD DETECTION LIMIT USING MTBE
REAGENT WATER ON PRIMARY DB-1 COLUMN
ANALYTE
Alachlor
Atrazine
Bromacil
Bromochl oroaceton i tr i 1 e
Bromodi chl oromethane
Bromoform
Carbon Tetrachloride
Chloral Hydrate
Chloropicrin
Chloroform
Cyanazine
Di bromoacetoni tr i 1 e
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1 , 2-Di bromoethane
Dichloroacetonitrile
1 , 1-Di chl oro-2-propanone
Endrin
Endrin Aldehyde
Endrin Ketone
Heptachlor
Heptachlor Epoxide
Hexachlorobenzene
Hexachl orocycl opentadi ene
Lindane (g-BHC)
Methoxychlor
Metolachlor
Metribuzin
Simazine
Fort.
Cone. ,
fjg/i
0.327
0.633
0.094
0.010
0.010
0.010
0.010
0.025
0.010
0.050
0.567
0.010
0.010
0.010
0.010
0.010
0.010
0.016
0.022
0.016
0.047
0.044
0.006
0.019
0.009
0.063
0.219
0.062
0.625
Obs£r.a
Cone. ,
fjg/i
0.384
0.764
0.099
0.011
0.012
0.018
0.011
0.029
0.009
0.054
0.757
0.016
0.011
0.020
0.020
0.009
0.0)11
0.023
0.023
0.016
0.062
0.050
0.006
0.019
0.015
0.057
0.254
0.100
0.794
Avg.
%Rec.
117
121
105
110
120
180
110
116
90
108
134
160
110
200
200
90
110
144
105
100
132
114
100
100
167
90
116
161
127
% RSD
2.13
3.56
10.05
5.42
7.50
8.12
6.32
5.61
7.65
34.04
13.93
12.78
4.55
15.15
12.54
4.28
6.22
2.57
2.25
5.14
43.65
1.64
5.44
31.81
9.89
4.85
3.20
12.45
5.95
MDLb
//g/L
0.025
0.082
0.030
0.002
0.003
0.004
0.002
0.005
0.002
0.055
0.316
0.006
0.001
0.009
0.008
0.001
0.002
0.002
0.002
0.002
0.081
0.002
0.001
0.018
0.004
0.008
0.024
0.037
0.142
EDLC
//g/L
0.500
0.324
0.055
0.009
0.005
0.006
0.004
0.011
0.014
0.075
0.685
0.010
0.007
0.009
0.008
0.005.
0.007
0.011
0.010
0.020
0.081
0.030
0.006
0.022
0.016
0.046
0.146
0.037
0.431
      551.1-38

-------
           TABLE 2.A.  METHOD DETECTION LIMIT USING MTBE (cont'd)
            NHAC1  PRESERVED REAGENT HATER ON PRIMARY DB-1 COLUMN
ANALYTE
Tetrachl oroethyl ene
Tri chl oroacetoni tri 1 e
1,1, 1-Tri chl oroethane
1,1, 2-Tri chl oroethane
Tri chl oroethyl ene
1,2,3-Trichloropropane
1,1, 1-Tri chl oro-2-propanone
Trifluralin
Surrogate ===>
Decafl uorobyphenyl
Fort.
Cone. ,
^g/L
0.
0.
0.
0.
0.
0.
0.
0.
10.
010
010
010
140
010
156
010
022
0
(a) Based upon the analysis of eight
(b) MDL designates the statistically
Obser.a
Cone. ,
//g/L
0.012
0.010
0.013
0.124
0.008
0.137
0.027
0.026
10.8
Avg.
%Rec.
120
100
130
89
80
88
270
118
108
replicate MTBE
derived MDL and
%
5
5
12
3
8
1
20
3
2
RSD
.04
.31
.35
.27
.68
.95
.53
.89
.38
sample
is cal

0
0
0
0
0
0
0
0

MDLb
//g/L
.002
.002
.005
.012
.002
.008
.016
.003

extracts
culated

0
0
0
0
0
0
0
0

by
EDLC
.004
.004
.005
.040 ;
.008
.028
.016
.010


     multiplying the standard deviation of the eight replicates by the
     student's t-value (2.998) appropriate for a 99% confidence level  and a
     standard deviation estimate with n-1 degrees of freedom.
(c)   Estimated Detection Limit (EDL) — Defined as either the MDL or a level
     of compound in a sample yielding a peak in the final extract with a
     signal  to noise (S/N) ratio of approximately 5, whichever is greater.
                                  551.1-39

-------
        TABLE 2.B.  METHOD DETECTION LIMIT USING MTBE
NH,C1  PRESERVED REAGENT HATER ON CONFIRMATION Rtx-1301
COLUMN
ANALYTE
Alachlor
Bromacil
Bromochl oroacetoni tri 1 e
Bromodi chl oromethane
Bromoform
Carbon Tetrachloride
Chloropicrin
Chloroform
Cyanazine
Di bromoacetoni tri 1 e
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1 , 2-Di bromoethane
Di chl oroacetoni tri le
l,l-Dichloro-2-propanone
Endrin
Endrin Aldehyde
Endrin Ketone
Heptachlor
Heptachlor Epoxide
Hexachl orobenzene
Hexachl orocycl opentadi ene
Lindane (g-BHC)
Methoxychlor
Metolachlor
Metribuzin
Simazine/Atrazine
Tetrachl oroethyl ene
Tri chl oroacetoni tri 1 e

0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
1
0
0
Fort.
Cone. ,
//g/L
.109
.094
.010
.010
.010
.010
.010
.010
.189
.010
.010
.010
.010
.010
.010
.016
.022
.047
.016
.044
.006
.019
.009
.188
.219
.062
.26 e
.010
.010
Obser.3
Cone. ,
0g/L
0
0
0
0
0
0

0
0
0
0
0
0
0
0
0
0
,0
0
0
0

0
0
0
0
1
0
0
.107
.134
.008
.012
.015
.0.11
NR d
.059
.279
.010
.021
.020
.039
.010
.009
.0,25
.034
.049
.018
.079
.006
NR
.011
.221
.280
.(#6
.619
.012
.006
Avg.
%Rec.
98
143
80
120
150
110
NR
590
148
100
210
200
390
100
90
156
155
104
113
180
100
NR
122
118
128
123
129
120
60
%RSD
1.
11.
9.
4.
29.
18.
70
65
49
34
51
70
MDLb
fjg/i
0
0
0
0
0
0
NR
2.
7.
4.
29.
9.
6.
4.
11.
4.
22.
5.
3.
84.
16.
82
56
87
30
95
44
11
65
09
45
49
79
71
47
0
0
0
0
0
0
0
0
0
0
0
0
0
0
.005
.047
.002
.002
.013
.006
NR
.005
.063
.001
.018
.006
.007
.001
.003
.003
.023
.008
.002
.202
.003
NR
6.
3.
1.
2.
2.
6.
16.
09
53
45
17
48
97
01
0
0
0
0
0
0
0
.002
.023
.012
.005
.121
.002
.003
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
d
0
0
0
0
0
0
0
EDL°
/>g/L
.076
.071
.015
.006
.013
.006
.062
.008
.065
.007
.018
.024
.007
.003
.015
.015
.030
.047
.010
.202
.011
.327 .
.009
.041
.268
.013
.629
.003
.010
                          551.1-40

-------
           TABLE 2.B.   METHOD DETECTION LIMIT USING MTBE (cont'd)
ANALYTE
1,1, 1-Tri chl oroethane
1,1, 2-Tri chl oroethane
Trichloroethylene
1,2, 3-Tri chl oropropane
1,1, 1-Tri chl oro-2-propanone
Trifluralin
Fort.
Cone. ,
y^g/L
0
0
0
0
0
0
.010
.140
.010
.156
.010
.022
Obser.a
Cone. ,
/jg/i
0
0
0
0
0
0
.020
.133
.009
.160
.011
.024
Avg.
%Rec .
200
95
90
103
110
109
%RSD
19
3
13
3
7
3
.22
.40
.77
.11
.11
.07
MDLb
/vg/L
0
0
0
0
0
0
.012
.014
.004
.015
.002
.002
EDLC
0.
0.
0.
0.
0.
0.
012
020
007
114
010
006
Surrogate ===>              10.0      10.6      106    1.78
Decaf1uorobyphenyl	


(a)  Based upon the analysis of eight replicate MTBE sample extracts. ;
(b)  MDL designates the statistically derived MDL and is calculated by
     multiplying the standard deviation of the eight replicates by the
     student's t-value (2.998) appropriate for a 99% confidence level and a
     standard deviation estimate with n-1 degrees of freedom.
(c)  Estimated Detection Limit (EDL) — Defined as either the  MDL or a
     level of compound in a sample yielding a peak in the final extract
     with a signal to noise (S/N) ratio of approximately 5, whichever is
   .  greater.                                                        .
(d)  NR indicates Not Reported since their was no peak detected for the
     eight replicate MDL determination.
(e)  The concentration of atrazine and simazine were added together for
     this determination since these two peaks coelute on the confirmation
     column.
                                 551.1-41

-------
      TABLE 3.A.  PRECISION.AND ACCURACY RESULTS USING MTBEa
NH,C1  PRESERVED FORTIFIED REAGENT WATER ON THE PRIMARY DB-1 COLUMN
ANALYTE
Alachlor
Atrazine
Bromacil
Bromochl oroacetoni tri 1 e
Bromodichloromethane
Bromoform
Carbon Tetrachloride
Chloropicrin
Chloroform
Cyanazine
Di bromoacetoni tri 1 e
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1 , 2-Di bromoethane
Dichl oroacetoni tri le
1 , 1-Di chl oro-2-propanone
Endrin
Endrin Aldehyde
Endrin Ketone
Heptachlor
Heptachlor Epoxide
Hexachlorobenzene
Hexachl orocycl opentadi ene
Lindane (g-BHC)
Methoxychl or
Hetolachlor
Metribuzin
Simazine
Tetrachl oroethyl ene
Tri chl oroacetoni tri le
Fortified ; Mean Meas.
Cone., /yg/L Cone., /yg/L
2.18
12.6
1.88
5.00
5.00 ''•
5.00
5.00 ;
5.00
5.00
3.77
5.00
5.00
5.00
5.00 ;
5.00
5.00
0.311
0.437
0.310 :
0.313
0.875
t
0.124
0.374
0.188
1.26
4.39
1.24 :
12.5
5.00
5.00
2.40
12.4
1.85
5.69
4.94
5.07
5.07
5.32
5.10
3.89
5.78
4.87
5.11
4.96
5.35
5.08
0.337
0.503
0.319
0.351
0.968
0.137
0.368
0.199
1.48
4.89
1.21
13.1
5.07
5.73
%RSD
1.47
1.71
3.13
0.71
1.14
0.72
1.72
1.38
1.30
2.85
1.43
0.71
0.59
0.73
0.57
0.72
1.40
1.32
1.52
2.84
0.65
0.89
1.18
1.41
2.84
0.87
3.94
2.02
1.62
1.34
Percent
Recovery
110
98
98
114
99
101
101
106
102
103
116
97
102
99
107
102
108
115
103
112
111
110
98
106
117
111
97
105
101
115
                            551.1-42

-------
       TABLE 3.A.   PRECISION AND ACCURACY  RESULTS  USING  MTBE" (cont'd)
      NH.C1  PRESERVED FORTIFIED REAGENT WATER ON  THE PRIMARY DB-1  COLUMN
Fortified Mean Meas.
ANALYTE Cone., //g/L Cone., UQ/i
1,1, 1-Trichl oroethane
1, 1,2-Trichloroethane
Trichloroethylene
1,2,3-Trichloropropane
1,1, 1-Tri chl oro-2-propanone
Trifluralin
5.00
2.80
5.00
3.12
5.00
0.439
5.02
2.92
4.87
3.08
5.30
0.503
%RSD
1.22
0.91
1.48
0.62
0.81
1.09 -
Percent
Recovery
100
104
97
99
106
115
Surrogate ===>
Decaf!uprobyphenyl
10.0
10.4
1.93
                                     104
(a)  Based upon the analysis of eight replicate MTBE sample extracts.
                                 551.1-43

-------
           TABLE 3.B.  PRECISION AND ACCURACY RESULTS USING MTBE a
     Na,SO, PRESERVED FORTIFIED REAGENT WATER ON THE PRIMARY DB-1 COLUMN
Fortified Mean Meas.
ANALYTE ' Cone., /jg/l Cone., fjg/i
Bromodi chl oromethane
Bromoform
Carbon Tetrachloride
Chloral Hydrate
Chloroform
Dibromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1 , 2-Di bromoethane
Tetrachl oroethyl ene
1,1, 1-Tri chl oroethane
Tri chl oroethyl ene
5.00
5.00
5.00 :
1.00
5.00
5.00
5.00
5.00
5.00
5.00 ;
5.00 :
4.91
5.05
5.08
0.93
4.96
4.83
5.07
4.90
5.06
5.01
4.81
%RSD
1,49
1.32
2.24
1.81
1.71
1.43
1.04
1.02
2.53
2.11
2.21
Percent
Recovery
98
101
102
93
99
97
101
98
101
100
96
Surrogate ===>
Decafluorobyphenyl
                                 10.0          10.2        1.88       102


(a)   Based upon the analysis of eight replicate MTBE sample extracts.
                                  551.1-44

-------
   TABLE 3.C.  PRECISION AND ACCURACY RESULTS USING MTBEa
NH4C1  PRESERVED FORTIFIED REAGENT WATER ON THE CONFIRMATION
                      Rtx-1301 COLUMN
ANALYTE
Alachlor
Bromacil
Bromochl oroacetoni tri le
Bromodichloromethane
Bromoform
Carbon Tetrachloride
Chloropicrin
Chloroform ,
Cyanazine
Di bromoacetoni tr i 1 e
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1,2-Dibromoethane
Dichl oroacetoni tri le
1, l-Dichloro-2-propanone
Endrin
Endrin Aldehyde
Endrin Ketone
Heptachlor
Heptachlor Epoxide
Hexachl orobenzene
Hexachl orocycl opentadi ene
Lindane (g-BHC)
Methoxychlor
Metolachlor
Metribuzin
Simazine/Atrazine
Tetrachl oroethyl ene
Tri chl oroacetoni tri 1 e
Fortified
Cone., jjg/L
2.18
1.88
5.00
5.00
5.00
5.00
5.00
5.00
3.77
5.00
5.00
5.00
5.00
5.00
5.00
0.310
0.440
0.310
0.310
0.880
0.124
0.374 .
0.188
1.26
4.39
1.24
25.1 b
5.00
5.00
Mean Meas.
Cone., //g/L
2.26
1.77
5.59
4.92
5.04
4.90
5.24
5.05
3.90
5.47
5.04
5.12
5.09
5.30
4.94
0.335
0.490
0.317
0.349
0.978
0.135
0.474
0.205
1.42
4.57
1.29
30.0
4.93
5.48
%RSD
0.81
3.50
0.86
1.02
0.73
1.72
1.20
1.20
2.30
0.58
0.90
0.54
1.82
0.55
0.70
2.08
2.13
1.63
1.06
0.80
0.59
7.19
0.75
2.30
3.43
1.15
1.11
1.65
1.31
. Percent
Recovery
104
94
112
98
101
98
105
101
103
109
101
102
102
106
99
108
111
102
113
111
109
127
109
113
104
104
119
99
110
                         551.1-45

-------
TABLE 3.C.  PRECISION AND ACCURACY RESULTS USING MTBE a (cont'd)
  NH,C1  PRESERVED FORTIFIED REAGENT WATER ON THE CONFIRMATION
                        Rtx-1301 COLUMN
ANALYTE
1,1, 1-Tri chl oroethane
1,1, 2-Tri chl oroethane
Trichloroethylene
1,2,3-Trichloropropane
1,1, 1-Tri chl oro-2-propanone
Trifluralin
Fortified
Cone. , //g/L
5.00
2.80
5.00
3.12
5.00
0.440
Mean Meas.
Cone., fjg/L
4.87
2.76
4.87
3.07
4.90
0.486
%RSD
1.66
1.52
1.52
0.88
0.89
0.93
Percent
Recovery
97
98
97 .
98
98
110
                                                    1.96
106
Surrogate -==>                   10.0         10.6
Decafluorobyphenyl
(a)  Based upon the analysis of eight replicate MTBE sample extracts.
(b)  Simazine and atrazine coelute on the confirmation column and therefore
     there results were added together.
                             551.1-46

-------
           TABLE 3.D.  PRECISION AND ACCURACY RESULTS USING MTBE a
        Na2S03 PRESERVED FORTIFIED REAGENT WATER ON THE CONFIRMATION
                               Rtx-1301  COLUMN
ANALYTE
Bromodi chl oromethane
Bromoform
Carbon Tetrachloride
Chloroform
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1,2-Dibromoethane
Tetrachl oroethyl ene
1,1,, 1-Trichloroethane
Trichl oroethyl ene
Fortified
Cone. , fjg/L
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
Mean Meas.
Cone. , fjg/L
4.88
5.03
4.90
4.90
5.15
5.07
5.02
4.89
4.84
4.83
%RSD
1.53
1.19
2.27
1.58
1.78
0.94
0.82
2.47
2.18
2.06
Percent
Recovery
98
101
98
98
103
101
100
98
97
97
Surrogate ===>                   10iO         10.3        1-.64
Decafluorobyphenyl
(a)  Based upon the analysis of eight replicate MTBE sample extracts.
103
                                  551.1-47

-------
       TABLE 4.A.   PRECISION AND ACCURACY RESULTS USING MTBE3
NH,.C1  PRESERVED FORTIFIED REAGENT HATER ON THE PRIMARY DB-1 COLUMN
ANALYTE
Alachlor
Atrazine
Bromacil
Bromochl oroaceton i tri 1 e
Bromodi chl oromethane
Bromoform
Carbon Tetrachloride
Chloropicrin
Chloroform
Cyanazine
Di bromoacetoni tr i 1 e
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1,2-Dibromoethane
Di chl oroacetoni tri 1 e
1 , 1-Di chl oro-2-propanone
Endrin
Endrin Aldehyde
Endrin Ketone
Heptachlor
Heptachlor Epoxide
Hexachlorobenzene
Hexachlorocyclopentadiene
Lindane (g-BHC)
Hethoxychlor
Metolachlor
Hetribuzin
Simazine
Tetrachl oroethyl ene
Tri chl oroaceton i tr i 1 e
Fortified
Cone., fjg/l
0.436
2.520
0.376
0.250
0.250
0.250
0.250
0.250
0.250
0.754
0.250
0.250
0.250
0.250
0.250
0.250
0.062
0.087
0.062
0.063
0.175
0.025
0.075
0.038
0.252
0.878
0.248
2.500
0.250
0.250
Mean Meas.
Cone. , fjg/l
0.515
2.994
0.376
' 0.281
0.276
0.260
0.299
0.285
0.264
0.761
0.276
0.266
0.261
0.274
0.268
0.261
0.073
0.108
0.062
0.059
0.206
0.030
0.074
0.047
0.298
1.056
0.264
2.960
0.263
0.291
%RSD
1.84
1.95
3.32
1.57
1.42
1.62
1.60
2,03.
1.94
1.97
1.89
1.20
1.82
1.89
1.12
0.91
2.65
1.29
0.76
10.29
0.90
3.77
3.22
2.74
3.24
1.00
2.15
2.71
1.93
1.02
Percent
Recovery
118
119
100
113
110
104
120
- 114
105
101
110
106
104
110
107
105
117
123
100
93
118
120
99
125
118
120
107
118
105
116
                              551.1-48

-------
       TABLE 4.A.  PRECISION AND ACCURACY RESULTS USING MTBEa (cont'd)
     NH^Cl  PRESERVED FORTIFIED REAGENT WATER ON THE PRIMARY DB-1 COLUMN
ANALYTE
1,1, 1-Tr i chl oroethane
1,1, 2-Tri chl oroethane
Trichloroethylene
1,2,3-Trichloropropane
1, l,l-Tn'ch1oro-2-propanone.
Trifluralin
Fortified
Cone., //g/L
0.250
0,560
0.250
0.624
0.250
0.088
Mean Meas.
Cone., //g/L
0.291
0.531
0.252
0.595
0.286
0.106
%RSD
3.65
0.85
1.20
0.83
3.72
1.50
Percent
Recovery
116
95
101
95
114
121
Surrogate ===>
Decaf1uorobyphenyl
                                 10.0      -    10.9        2.49       109
                                                              s.

(a)   Based upon the analysis of eight replicate MTBE sample extracts.
                                  551.1-49

-------
           TABLE 4.B.  PRECISION AND ACCURACY RESULTS  USING  MTBE8
     Na,SO, PRESERVED FORTIFIED REAGENT WATER ON THE PRIMARY  DB-1  COLUMN
ANALYTE
Bromodi chl oromethane
Bromoform
Carbon Tetrachloride
Chloral Hydrate
Chloroform
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1,2-Dibromoethane
Tetrachl oroethyl ene
1,1, 1-Tri chl oroethane
Tri chl oroethyl ene
Fortified
Cone., /jg/l
0.250
0.250
0.250
0.250
0.250
0.250
0.250
0.250
0.250
0.250
0.250
Mean Meas.
Cone. , //g/L
0,270
0.257
0.287
0.258
0.248
0.261
0.258
' 0.243
0.256
0.276
0.246
%RSD
1.77
2.04
5.18
4.12
1.88
1.36
1.26
0.90
1.95
5.72
1,01
Percent
Recovery
108
103
115
103
99
105
103
97
102
110
98
Surrogate -«>                   10.0          10.6         3.51        106
Decafluorobyphenyl
(a)  Based upon the analysis of eight replicate MTBE sample  extracts..
                                   551.1-50

-------
           TABLE  5.A.   PRECISION AND ACCURACY RESULTS USING MTBEa
NH4C1  PRESERVED FORTIFIED FULVIC ACID ENRICHED REAGENT WATER6 ON THE PRIMARY
                                 DB-1 COLUMN
ANALYTE
Alachlor
Atrazine
Bromacil
Bromochloroacetonitrile
Bromodi chl oromethane
Bromoform
Carbon Tetrachloride
Chloropicrin
Chloroform
Cyanazine
Dibromoacetonitrile
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1,2-Dibromoethane
Dichloroacetoni trile
l,l-Dichloro-2-propanone
Eridrin
Endrin Aldehyde
Endrin Ketone
Heptachlor
Heptachlor Epoxide
Hexachlorobenzene
Hexachlorocyclopentadiene
Lindane (g-BHC)
Methoxychlor
Metolachlor
Metribuzin
Simazine
Tetrachl oroethyl ene
Fortified
Cone. , jjq/L
2.18
12.6
1.88
1.00
1.00
1.00
1.00
1.00
1.00
3.77
1.00
1.00
.1.00
1.00
1.00
1.00
0.311
0.437
0.310
0.313
0.875
0.124
0.374
0.188
1.26
4.39
1.24
12.5
1.00
Mean Meas.
Cone., fjq/l
2.38
11.6
1.89
1.11
0.87
0.97
0.88
1.13
1.03
4.02
1.14
0.89
0.93
0.96
1.05
1.03
0.325
0.505
0.319
0.358
0.978
0.139
0.363
0.206
1.41
4.84
1.30
12.0
0.90
%RSD
1.57
2.31
3.33
1.51
1.93
1.50
3.91
2.49
2.47
3.99
1.61
1.78
1.37
1.58
0.98
0.90
3.50
1.99
2.62
5.45
1.28
1.82
3.55
1.79
4.78
1.27
2.08
1.09
4.02
Percent
Recovery
109
92
101
111
87
97
88
113
103
107
114
89
93
96 .
105
103
104
116
103
114
112
112
97
110
112
110
105
96
90
                                 551.1-51

-------
      TABLE 5.A.   PRECISION AND ACCURACY RESULTS USING MTBE a (cont'd)
NH,C1 PRESERVED FORTIFIED FULVIC ACID ENRICHED REAGENT WATER*5 ON THE PRIMARY
                                 DB-1  COLUMN
ANALYTE
Trichloroacetonitrile
1,1, 1-Tri chl oroethane
1 , 1 , 2-Tri chl oroethane
Trichloroethylene
1,2,3-Trichloropropane
1,1, 1-Tri chl oro-2-propanone
Trifluralin
Fortified
Cone., /yg/L
1.00
1.00
2.80
1.00
3.12
1.00
0.439
Mean Meas.
Cone. , //g/L
1.11
0.96
2.81
0.93
; 2.92
1.10
0.517
%RSD
2.41
3.89
2.89
3.55
0.82
2.05
1.27
Percent
Recovery
111
96
100
93
93
110
118
Surrogate ===>                     10.0         10.4       1.84      104
Decaf1uorobyphenyl
(a)  Based upon the analysis of eight  replicate MTBE sample extracts.
(b)  Reagent water fortified at 1.0 mg/L with fulvic acid extracted from
     Ohio River water.  Sample simulated high TOC matrix.
                                   551.1-52

-------
           TABLE 5.B.  PRECISION AND ACCURACY RESULTS USING MTBE a
 Na2S03 PRESERVED FORTIFIED  FULVIC ACID  ENRICHED  REAGENT WATER ON THE  PRIMARY
                                 DB-j.COLUMN
ANALYTE
Bromodi chl oromethane
Bromoform
Carbon Tetrachloride
Chloral Hydrate
Chloroform
Dibromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1,2-Dibromoethane
Tetrachl oroethyl ene
1 , 1 , 1-Trichl oroethane
Tri chl oroethyl ene
Fortified
Cone. , fjg/L
1.00
1.00
1.00
1.00
1.00
1.00
1.00
1.00
1.00
1.00
1.00
Mean Meas.
Cone., fjg/L
0.87
0.97
0.88
0.90
0.96
0.88
0.92
0.93
0.90
0.97
0.94
%RSD
1.13
1.28
1-71
0.95
1.51
1.25
0.98
1.01
2.07
1.57
1.62
Percent
Recovery
87
97
88
90
96
88
92
93
90
97
94
Surrogate ===>                     10.0         10.6       2.56      106
Decaf1uorobyphenyl

(a)  Based upon the analysis of eight replicate MTBE sample extracts.
(b)  Reagent water fortified at 1.0 mg/L with fulvic acid extracted from
     Ohio River water.  Sample simulated high TOC matrix.
                                  551.1-53

-------
TABLE 6.A.   PRECISION  AND  ACCURACY  RESULTS  USING MTBEa
NH,C1  PRESERVED FORTIFIED GROUND WATER6 ON THE PRIMARY
                      DB-1  COLUMN
ANALYTE
Alachlor
Atrazine
Bromacil
Bromochl oroacetoni tri 1 e
Bromodi chl oromethane
Bromoform
Carbon Tetrachloride
Chloropicrin
Chloroform
Cyanazine
Di bromoacetoni tri 1 e
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1,2-Dibromoethane
Di chl oroacetoni tri 1 e
1 , 1-Di chl oro-2-propanone
Endrin
Endrin Aldehyde
Endrin Ketone
Heptachlor
Heptachlor Epoxide
Hexachlorobenzene
Hexachl orocycl opentadi ene
Lindane (g-BHC)
Methoxychlor
Metolachlor
Metribuzin
Simazine
Tetrachl oroethyl ene
Unfort.
matrix
cone. ,
//g/L
ND c
N.D
ND
ND
1.70
20.1
ND
ND
0.571
ND
ND
6.00
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
Fort.
Cone. ,
/vg/L
8,72
50,4
7 , 52
5.00
5.00
5.00
5.00
5.00
5.00
15.1
5.00
5:.00
5,00
5.00
5.00
5.00
1.24
1.75
1.24
1.25
3.50
0 . 50
1.50
0.75
5.04
.17.6
4.96
50.0
5.00
Mean
Meas.
Cone. ,
//g/L
9.01
46.7
6.53
5.74
6.68
24.8
4.99
5.29
5.73
15.4
5.84
11.1
5.04
4.87
5.29
5.01
1.32
1.91
1.22
1.33
3.67
0.509
1.41
0.773
5.60
18.2
4.85
48.3
4.97
%RSD
2.93
3.30
7.81
1.38
2.59
1.61
6.65
3.59
3.68
6.07
1.59
1.89
1.64
1.90
1.52
1.30
4.81
2.36
3.77
4.46
2.92
3.42
3.70
1.91
5.86
3.06
6.15
3.30
6.29
Percent
Recovery
103
93
87
115
100
95
100
106
103
102
117
102
101
97
106
100
106
109
98
106
105
103
94
103
111
103
98
97
99
                        551.1-54

-------
       TABLE 6.A.  PRECISION AND ACCURACY RESULTS USING MTBEa (cont'd)
           NH4C1 PRESERVED FORTIFIED GROUND WATER6 ON THE PRIMARY
                                 DB-1 COLUMN



ANALYTE
Trichloroacetonitrile
1,1,1-Trichloroethane
1,1, 2-Tri chl oroethane
Trichloroethylene
1,2, 3-Tri chl oropropane
1,1, 1-Trichl oro-2-propanone
Trifluralin
Unfort.
matrix
cone.,
//9/L
ND
1.77
ND
ND
0.340
ND
ND

Fort.
Cone. ,
ywg/L
5.00
5.00
11.2
5.00
12.5
5.00
1.76
Mean
Meas.
Cone.,
//9/L
5.59
6.62
10.4
4.74
12.5
5.21
1.94



%RSD
4.89
4.60
2.98
5.78
3.92
1.58
3.38


Percent
Recovery
112
97
93
95
97
104
110
Surrogate ===>                           10.0       10.4      2.25   104
Decaf1uorobyphenyl

(a)  Based upon the analysis of eight replicate MTBE sample extracts.
(b)  Chlorinated ground water from a water source displaying a hardness of
     460 mg/L as CaC03.
(c)  ND indicates not detected above the EDL.
                                  551.1-55

-------
           TABLE 6.B.  PRECISION AND ACCURACY RESULTS USING MTBEa
     Na,SO, PRESERVED FORTIFIED GROUND WATER6 ON THE'PRIMARY'DB-1  COLUMN
ANALYTE
Bromodichloromethane
Bromoform
Carbon Tetrachloride
Chloral Hydrate
Chloroform
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1 , 2-Di bromoethane
Tetrachl oroethyl ene
1,1, 1-Tri chl oroethane
Trichl oroethyl ene
Unfort.
matrix
cone. ,
X/9/L
1.77
20.5
ND c
ND
0.600
6.16
ND
ND
ND
1.91
ND
Fort.
Cone. ,
/>g/L
5.00
5.00
5 . 00
2.00
5.00
5.00'
5.00
5.00
5.00
5.00
5.00
Mean
Meas.
Cone.,
//g/L
6.64 :
24.6
4.99
1.84,
5.22
11.0
5.01 .
4,79 :
4.95 '
6.73
4.69
%RSD
1.70
1.63
2.72
1,38
1.89,
1.53
1.19
1.86
2.49
3.18
2.38
Percent
Recovery
97
82
100
92,
92
98
100
96
99;
96
94











Surrogate —=>
Decaf1uorobyphenyl
10.0
10.1
                                                             8.71
101
(a)  Based upon the analysis of eight replicate MTBE sample extracts.
(b)  Chlorinated ground water from a water source displaying a hardness of
     460 mg/L as CaC03.
(c)  ND indicates Not Detected above the detection limit.
                                  551.1-56

-------
            TABLE 7.   LABORATORY PERFORMANCE CHECK SOLUTION
parameter
Instrument
Sensitivity
Chromatographic
Performance
Column
Performance

Analyte
Breakdown
Analyte
Lindane
(gamma-BHC)
Hexachl orocycl opentadi ene
Bromodichloromethane
Trichloroethylene
Bromacil
Alachlor
Endrin
Cone.,
jt/g/mL
in MTBE
or pentane
0.000200
0.0200
0.0300
0.0300
0.0830
0.0830
0.0300
Acceptance
Criteria
Detection of
Analyte; = . .-.
Signal to
Noise > 3
PGF between
0.80 and 1.1 5a
Resolution >
0.50b
Resolution >
0.50
%BDC < 20 %
      PGF  =  Peak Gaussian  Factor.  Calculated  using  the  equation-
            1.83 x  W(l/2)
      PGF  =      ------ - -------------
b
where W(l/2)  is the peak width at half height and W(l/10) is the
peak width at tenth height.

Resolution between the two peaks as defined by the equation:

R = -----
      W
where t is the difference in elution times between the two peaks and
W is the average peak width, at the baseline, of the two peaks.

%BD = Percent Breakdown.  Endrin breakdown calculated using the
equation.
            (AREA Endrin Ketone + AREA Endrin Aldehyde)
          =	x
          (AREA Endrin Ketone + AREA Endrin Aldehyde + AREA Endrin)
Note:     If laboratory EDL's differ from those listed in this method,
          concentrations of the LPC standard must be adjusted to be
          compatible with the laboratory EDL's.
                              551.1-57

-------
   TABLE 8.  METHOD DETECTION LIMIT USING PENTANE
NH4C1  PRESERVED REAGENT WATER ON PRIMARY DB-1 COLUMN
ANALYTE
Alachlor
Atrazine
Bromacil
Bromochl oroaceton i tri 1 e
Bromodi chl oromethane
Bromoform
Carbon Tetrachloride
Chloropicrin
Chloroform
Cyanazine
Di bromoacetoni tri 1 e
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1,2-Dibromomethane
Di chloroacetoni tri 1 e
i;i-Dichloro-2-propanone
Endrin
Endrin Aldehyde
Endrin Ketone
Heptachlor
Heptachlor Epoxide
Hexachlorobenzene
Hexachl oropentadi ene
Lindane (g-BHC)
Methoxychlor
Metol achl or
Metribuzin
Simazine
Tetrachl oroethyl ene
Fort.
Cone.
A/g/L
0.109
0.633
0.094
0.040
0.040
0,
0,
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
.040
.040
.040
.040
.189
.040
.040
.040
.040
.040
.040
.016
.022
.016
.016
.044
.0062
.040
.0094
.063
.219
.062
.625
.040
Observ.b
Cone.
A/g/L
0.095a
0.663
0.058
0.
,047
0.054
0,033
0,
.060
0,045
0,
0.
0
0
0
0
0
0
0
0
0
0.
0
0
0
0
0
0
0
0
0
,110
170a
,046
,050
.053
.053
.037
.042
.019
.023
•.014
Olla
.045
,008
.022
.006
.069
.267
.076
.662
.052
Avg.
%Rec.
87
105
62
118
135
83
150
113
275
90
115
125
133
133
93
105
119
105
88
69
102
129
55
64
110
122
123
106
130
%RSD
5.37
5.00
21.44
3.61
42.05
20.60
27.76
4.25
24.36
13.37
3
5
5
19
20
4
4
5
9
18
5
9
24
91
12
10
18
9
5
.84
.48
.39
.85
.09
.86
.69
.52 .
. 50
.14
.02
.56
.42
.20
.76
.35
.15
.42
.33
MDLC
A/g/L
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
015
099
037
005
068
020
050
006
080
068
005
008
009
032
022
006
,003
0.004
0.004
0.006
0.007
0.002
0.016
0.017
0
0
0
0
0
.026
.083
.041
.187
.008
EDLd
A/g/L
0.050
0.390
0.330
0.026
0.068
0.035
0.050
0.023
0.080
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
.200
.030
.026
.017
.032
.042
.022
.016
.022
.020
.009
.0*16
.002
.016
.017
.066
.172
.041
.420
.016
                       551.1-58

-------
          TABLE  8.   METHOD  DETECTION  LIMIT  USING  PENTANE  (cont'd)
             NH4C1 PRESERVED REAGENT WATER ON  PRIMARY DB-1 COLUMN
ANALYTE
Tri chl oroaceton itr i 1 e
1,1,1-Trichloroethane
1,1, 2-Trl chl oroethane
Trichloroethylene
1,2,3-Trichloropropane
1, 1, l-Trichloro-2-propanone
Trifluralin
Fort.
Cone.
//g/L
0.040
0.040
0.140
0.040
0.156
0.040
0.040
Observ.b
Cone.
//g/L
0.048
0.058
0.141
0.064
0.151
0.045
0.021
Avg.
%Rec.'
120
145
101
160
97
113
53
%RSD
2.79
4.26
4.01
21.80
3.54
3.65
19.28
MDLC
//g/L
0.004
0.007
0.017
0.042
0.016
0.005
0.012
EDLd
//g/L
0.014
0.017
0.052
0.042
0.116
0.024
0.012
Surrogate ===>
Decaf1uorobyphenyl
10.0
11.2   112
3.98
(a)  Quantitated from confirmation column due to.baseline interference on
     primary column.
(b)  Based upon the analysis of eight replicate pentane sample extracts.
(c)  MDL designates the statistically derived MDL and is calculated by  '
     multiplying the standard deviation of the eight replicates by the
     student's t-value (2.998) appropriate for a 99% confidence level  and a
     standard deviation estimate with n-1 degrees of freedom.
(d)  Estimated Detection Limit (EDL)  — Defined as either the  MDL or a level
     of compound in a sample yielding a peak in the final  extract with a
     signal  to noise (S/N)  ratio of approximately 5,  whichever is greater
                                 551.1-59

-------
            TABLE 9.  PRECISION AND ACCURACY RESULTS3
                          USING PENTANE
NH4C1  PRESERVED FORTIFIED REAGENT HATER ON THE PRIMARY DB-1
COLUMN
ANALYTE
Alachlor
Atrazine
Bromacil
Bromochl oroacetoni tri 1 e
Bromodichloromethane
Bromoform
Carbon Tetrachloride
Chloropicrin
Chloroform
Cyanazine
Dibromoacetonitrile
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
1,2-Dibromoethane
Dichloroacetonitrile
l,l-Dichloro-2-propanone
Endrin
Endrin Aldehyde
Endrin Ketone
Heptachlor Epoxide
Heptachlor
Hexachlorobenzene
Hexachl orocycl opentadi ene
Lindane (g-BHC)
Methoxychlor
Metolachlor
Hetribuzin
Simazine
Tetrachl oroethyl ene

Fortified
Cone. , jjg/L
2.18
12.6
1.88
5.00
5.00 :
5.00
5.00
5.00
5.00
3.77
5.00
5.00
5.00
5.00
5.00
5.00
0.311
0.437
0.310':
0.875
0.313b'
0.124
0.374
0.188.
1.26
4.39 :
1.24
12.5 '
5.00 [
551.1-60;
Mean Meas.
Cone. , //g/L
1.98 b
12.0
1.74
4.63
4.46
4.81
4.61
4.51
4.95
4.00 b
4.80
4.23
4.73
4.69
4.73
4.78
0.312
0.443
0.311
0.866
0.30
0.123
0.384
0.176
1.28
4.42
1.34
12.5
4.46

%RSD
5.09
3.09
2.95
3.18
4.07
2.76
4.14
2.46
2.90
2.59
2.87
3.38
3.00
2.54
3.39
3.04
2.61
2.29
2.10
2.11
3.47
2.51
3.30
10.23
3.03
2.36
2.13
2.20
3.67

Percent
Recovery
91
95
93
93
89
96
92
90
99
106
96
85
95
94
95
96
100
101
100
99
97
99
103
94
102
101
108
100
89


-------
            TABLE 9.  PRECISION AND ACCURACY RESULTS3 (cont'd)
                              USING PENTANE
    NH4C1  PRESERVED FORTIFIED REAGENT WATER ON THE PRIMARY DB-1  COLUMN
ANALYTE
Trichloroacetonitrile
1, 1 , 1-Trichloroethane
1, 1,2-Trichloroethane
Trichloroethylene
1,2,3-Trichloropropane
1 , 1 , 1-Tri chl oro-2-propanone
Trifluralin
Surrogate===>
Decaf luorobyphenyl
Fortified
Cone., fjg/l
5.00
5.00
2.80
5.00
3.12
5.00
0.439
10.0
Mean Meas.
Cone., jjq/l
5.07
4.70
2.62
4.84
3.13
4.88
0.446
10.7
%RSD
4.02
3.39
2.03
2.98
1.76
2.80
2.74
1.88
Percent
Recovery
101
94
93
97
100
98
102
107
(a)      Based upon the analysis of eight replicate pentane sample extracts.

(b)      Quantitated from confirmation column due to baseline interference
        on primary column.
                                551.1-61

-------
TABLE 10.  ANALYTE PEAK IDENTIFICATION, RETENTION TIMES,
  CONCENTRATIONS AND  CONDITIONS USING MTBE FOR FIGURE 1
     NH,C1  PRESERVED  FORTIFIED  REAGENT WATER  ON THE
                   PRIMARY DB-1 COLUMN
PEAK
#
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
Retention
Time3
ANALYTE minutes
Chloroform
1,1,1-Trichloroethane
Carbon Tetrachloride
Trichloroacetonitrile ;
Dichloroacetonitrile
Bromodi chl oromethane
Trichloroethylene
Chloral Hydrate
1 , 1-Di chl oro-2-Propanone
1 , 1 , 2-Tri chl oroethane
Chloropicrin
Di bromochl oromethane
Bromochl oroacetoni tri 1 e
1,2-Dibromoethane (EDB)
Tetrachl oroethyl ene
1 , 1 , 1-Tri chl oropropanone
Bromoform
Di bromoacetoni tri 1 e
1,2,3-Trichloropropane
1 , 2-Di bromo-3-chl oropropane (DBCP)
Surrogate: Decaf luorobiphenyl
Hexachl orocycl opentadi ene
Trifluralin
Simazine
Atrazine
Hexachl orobenzene
Lindane (gamma-BHC)
Metribuzin
7.04
8.64.
9.94
10.39
12.01
12.42
12.61
13.41
14.96
19.91
23.10
23.69
24.03
24.56
26.24
27.55
29.17
29.42
30.40
35.28
36.35
40.33
45.17
46.27
46.55
47.39
47.95
50.25
Cone.
' M/l
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
44.8
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
50.0
5.00
10.0
28.0
7.04
200
200
1.98
30.1
19.9
                         551.1-6?.

-------
           TABLE  10.  ANALYTE  PEAK IDENTIFICATION,  RETENTION TIMES,
        CONCENTRATIONS AND  CONDITIONS  USING MTBE FOR FIGURE 1 (cont'd)
                NH4C1  PRESERVED FORTIFIED REAGENT WATER ON THE
                              PRIMARY DB-1 COLUMN
PEAK
#
29
30
31
32
33
34
35
36
37
38
NOTE:
ANALYTE
Bromacil
Alachlor
Cyanazine
Heptachlor
Metolachlor
Heptachlor Epoxide
Endrin
Endrin Aldehyde
Endrin Ketone
Methoxychlor
Bromofiuorobenzene (ret.
standard was not included
Retention
Time3
minutes
52.09
52.25
53.43
53.72
55.44
58.42
64.15
65.46
72.33
73.53
time 31.00 rri'in.) as the
in this chromatogram.
Cone.
ywg/L
30.1
34.9
60.4
5.00
70.0
14.0
5.00
7.00
4.96
20.1
internal
(a)   Column A -
     0.25 mm  ID x 30 m fused  silica  capillary  with  chemically
     bonded methyl polysiloxane phase  (J&W,  DB-1, 1.0  urn  film
     thickness or equivalent).  The  linear velocity of the
     helium carrier is established at  25 cm/sec  at  35°C.
The column oven is temperature programmed as follows:
[1]  HOLD at 35°C for 22 min
[2]  INCREASE to 145°C at 10°C/min and  hold  at  145°C for 2 min.
[3]  INCREASE to 225°C at 20°C/min and  hold  at  225°C for 15
     min.
[4]  INCREASE to 260°C at 10°C/min and  hold  at  260°C for 30
     min. or until all expected compounds have eluted.
Injector temperature:  200°C
Detector temperature:  290°C
                                   551.1-63

-------
FIGURE 1.  FORTIFIED  REAGENT WATER EXTRACT USING MTBE ON PRIMARY DB-1 COLUMN
                                               15
                                           12
                                    _A	

                                                   17


                                                 1«  "18
                                                      "
                                                             H
                                                                      "
                                             .... -._! _ , _ . - . - . - 1 - , - . - . - . - 1
                    IB       15
2i       25

  MINUTES
                                                     38
                              3*
                 21
        23
          25
         '(
         bUL
                         33

                 li
                                         i  .... _.i
                                                      35       48
45       Si       55
                                 69
     65

  MINUTES
                                          78       75       60        85
                                    551.1-64

-------
TABLE 11.  ANALYTE PEAK IDENTIFICATION, RETENTION TIMES,
  CONCENTRATIONS  AND  CONDITIONS  USING  MTBE  FOR  FIGURE  2
     NH4C1  PRESERVED  FORTIFIED REAGENT WATER ON THE
                  CONFIRMATION RtX-1301
PEAK
.#
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19 .
20
21
22
23
24
25
26
27
ANALYTE
Chloroform
1,1,1-Trichloroethane
Carbon Tetrachloride
Trichloroacetonitrile
Trichloroethylene
Bromodichloromethane
1 , l-Dichloro-2-Propanone
Chloropicrin
Tetrachl oroethyl ene
1 , 1 , 2-Tri chl oroethane
Dichloroacetonitrile
Di bromochl oromethane
1,2-Dibromoethane (EDB)
1,1, 1-Tri chl oropropancne
Bromochl oroacetoni tri 1 e
Bromoform
1 , 2 , 3-Tri chl oropropane
Dibromoacetonitrile
1 , 2-Di bromo-3-chl oropropane (DBCP)
Surrogate: Decaf luorobiphenyl
Hexachl orocycl opentadi ene
Trifluralin
Hexachl orobenzene
Atrazine/Simazine
Lindane (gamma-BHC)
Heptachlor
Metribuzin
Retention
Time3
minutes
7.73
7.99
8.36
10.35
11.96
15.28
20.50
23.69
24.77
25.01
25.21
26.32
26.46
28.47
29.86
30.36
31.73
32.77
36.11
36.28
39.53
45.43
46.47
48.56
49.68
53.15
53.92
Cone.
//g/L
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
44.8
5.00
5.00
5.00
5.00.
5.00
5.00
50.0
5.00
5.00
10.0
28.0
7.04
1.98
400
30.1
5.00
19.9
                       551.1-65

-------
           TABLE 11.   ANALYTE PEAK IDENTIFICATION,  RETENTION TIMES,
        CONCENTRATIONS AND CONDITIONS USING MTBE FOR FIGURE 2 (cont'd)
                NH,C1 PRESERVED FORTIFIED REAGENT WATER ON THE
                 	CONFIRMATION RtX-1301	

                                                 Retention
                                                   Time3         Cone.
           ANALYTE      	        minutes       /vg/L

     28    Alachlor                                   54.38       34.9

     29    Metolachlor                               57.07       70.0

     30    Heptachlor Epoxide                        59.05       14.0

     31    Bromacil                                   59.60       30.1

     32    Cyanazine                             -    59.89       60.4

     33    Endrin                                     65.24        5.00

     34    Endrin Aldehyde                           71.56        7.00

     35    Methoxychlor                              76.73       20.1

     36    Endrj n Ketone	81.28	4.96	
    "NOTE':	BTbmoTTuorobenz'erie (ret. time 31.30 min.) as the internal
               standard was not included in this chromatogram.

(a)   Column B -     0.25 mm ID x 30 m with chemically bonded 6 %
                    cyanopropylphenyl / 94 % dimethyl polysiloxane phase
                    (Restek, Rtx-1301, 1.0 fim film thickness or equivalent).
                    The linear velocity of the helium carrier gas is
                    established at 25 cm/sec at 35°C.
               The column oven is temperature programmed as follows:
               The column oven is temperature programmed as follows:
                [1]  HOLD at 35°C for 22 min
                [2]  INCREASE to 145°C at lp°C/min  and hold at 145°C  for  2 min
                [3]  INCREASE to 225°C at 20°C/min  and hold at 225°C  for  15  min
                [4]  INCREASE to 260°C at 10°C/min  and hold at 260°C  for  30
                    min. or until all expected compounds have eluted.
                Injector temperature:  200°t
               Detector temperature:  290°C
                                    551.1-66

-------
FIGURE 2.   FORTIFIED REAGENT WATER EXTRACT USING  MTBE ON  CONFIRMATION Rtx-1301
COLUMN

                                          I     8
                                                     12
                                                         U
                                                      ,13
                                                           15
                                                             16
                                                               17
                                                                        U
                                                                             21
             -J—.—i—i—i—I—i—i—.—i—i—,	1—.—i—i—,—,	,	i,  ....  i
              5         10        15
2B        25
  MINUTES
                                                           3B        35
                                   38
                         27
         22
                               29
                          28
                                    32
                                              33
                                              A
                 31
                                                                   35
        -1	i	'	1	1	1	.	.	1	.	1	1	L	:	,	1	,	,	,__,	I	,	.  .  .  I
        45        50        55
60        65
      MINUTES
             78        75        81         85
                                       551.1-67

-------
TABLE 12.  ANALYTE PEAK IDENTIFICATION, RETENTION TIMES, CONCENTRATIONS
               AND CONDITIONS USING PENTANE FOR FIGURE 3
            NH4C1 PRESERVED FORTIFIED REAGENT WATER ON THE
                          PRIMARY DB-1 COLUMN
Retention
PEAK Time3
-# ANALYTE minutes
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
Chloroform
1 , 1 , 1-Tri chl oroethane
Carbon Tetrachloride
Trichloroacetonitrile
Di chl oroaceton i tri 1 e
Bromodi chl oromethane
Tri chl oroethyl ene
1 , 1-Di chl oro-2-Propanone
1 , 1 , 2-Tri chl oroethane
Chloropicrin
Di bromochl oromethane
Bromochl oroacetoni tri le
1,2-Dibromoethane (EDB)
Tetrachl oroethyl ene
1 , 1 , 1-Tri chl oropropanone
Bromoform
Di bromoacetoni tri 1 e
1,2, 3-Tri chl oropropane
Internal Standard: Bromofluorobenzene
1 , 2-Di bromo-3-chl oropropane (DBCP)
Surrogate: Decaf luorobiphenyl
Hexachl orocycl opentadi ene
TrifluraTin
Simazine
Atrazine
Hexachl orobenzene
Lindane (gamma-BHC)
8.41
10.26
11.56
12.03
13.53
13.73
13.89
15.60
18.57
20.49
21.03
21.25
22.03
24.75
27.94
30.97
31.45
32.82
33.60
38.34
39.48
43.92
49.04
50.08
50.37
51.11
51.66
Cone.
fJ<3/l
5.00
5.00
5.00
5.00
5.00
5.00
5.00
5.00
44.8
5.00
5.00
5.00
5.00
5.00
5.00
5.00
. 5.00
50.0
1.00 /yg/mL in
pentane extract
5.00
10.0
28.0
7.04
200
200
1.98
30.1
                                551.1-68

-------
TABLE 12.  ANALYTE PEAK  IDENTIFICATION,  RETENTION TIMES,  CONCENTRATIONS
           AND CONDITIONS USING PENTANE FOR FIGURE 3 (cont'd)
            NH4C1 PRESERVED FORTIFIED REAGENT WATER ON THE  .
                          PRIMARY DB-1 COLUMN
PEAK
#
28
29
30
31
32
33
34
35
36
37
38
(a) Col
ANALYTE
Metribuzin
Bromacil
Alachlor
Cyanazine
Heptachlor
Metolachlor
Heptachlor Epoxide
Endrin
Endrin Aldehyde
Endrin Ketone
Methoxychlor
umn A - 0.25 mm
Retention
Time3
minutes
53.95
55.72
55.87
57.04
57.21
59.13
62.50
68.00
69.25
75.74
76.98
ID x 30 m fused silica capillary
Cone.
0g/L
19.9
30.1
34.9
60.4
5.00
70.0
14.0
5.00
7.00
4.96
20.1
r with chemically
                bonded methyl  polysiloxane .phase (J&W,  DB-1,  1.0 fim film
                thickness  or  equivalent).  The linear velocity of the
                helium carrier is  established  at 25 cm/sec at 35°C.
           The column oven  is  temperature programmed as follows:
           [1]  HOLD at  15°C for 0 min
           [2]  INCREASE to 50°C at 2°C/min  and hold at 50°C for 10 min
           [3]  INCREASE to 225°C at 10°C/min and hold at 225°C for 15 min
           [4]  INCREASE to 260°C at 10°C/min and hold at 260°C for 30
                min. or  until  all  expected compounds have eluted.
           Injector temperature:   200°C
           Detector temperature:   290°C
                               551.1-69

-------
   FIGURE 3.
COLUMN
FORTIFIED REAGENT WATER EXTRACT USING PENTANE ON  PRIMARY DB-1
                                                  U
                                                                           21
                             5 J
                                          IB
                                              13
                                                        15
                                                               17   19
                                                                             21
                                                             1	,	.	1	1
                       11       15
                         2B        25
                           MINUTES
                                                            30        35       M
                          28
                 23
   KM

  Nl (27
                                31
              3i  32
                V  33
                                                      36
                                                    35
                                                                     38
                                                                  37
          45        5B       55
                               65        70
                            MINUTES
                                                                 75        80        15
                                         551.1-70

-------
                                                                                         I
METHOD.552.2   DETERMINATION OF HALOACETIC ACIDS AND DALAPON  IN DRINKING WATER
               BY LIQUID-LIQUID EXTRACTION, DERIVATIZATION AND GAS
               CHROMATOGRAPHY WITH ELECTRON CAPTURE DETECTION.
                                 Revision 1.0
J.W. Hodgeson (USEPA), J. Collins and  R.E. Barth (Technology Applications
Inc.) - Method 552.0, (1990)

J.W. Hodgeson (USEPA), D. Becker (Technology Applications Inc.) -? Method
552.1, (1992)

D.J. Munch, J.W. Munch (USEPA) and A.M. Pawlecki (International Consultants,
Inc.), Method 552.2, Rev. 1.0, (1995)
                     NATIONAL EXPOSURE RESEARCH LABORATORY
                      OFFICE OF RESEARCH AND DEVELOPMENT
                     U.S. ENVIRONMENTAL PROTECTION AGENCY
                            CINCINNATI, OHIO 45268
                                    552.2-1

-------
          METHOD 552.2  DETERMINATION OF HALOACETIC  ACIDS  AND  DALAPON
         IN DRINKING WATER BY LIQUID-LIQUID EXTRACTION,  DERIVATIZATION
            AND GAS CHROMATOGRAPHY WITH ELECTRON CAPTURE DETECTION
1.   SCOPE AND APPLICATION

     1.1  This is a gas chromatographic (GC) method (1-8) applicable to the
          determination of the listed halogenated acetic acids in drinking
          water, ground water, raw source water and water at any intermediate
          treatment stage.  In addition, the chlorinated herbicide, Dalapon,
          may be determined using this method.

                                            Chemical Abstract Services
               Analvte                             Registry Number

          Bromochloroacetic Acid (BCAA)                5589-96-3
          Bromodichloroacetic Acid (BDCAA)            7113-314-7
          Chlorodibromoacetic Acid (CDBAA)             5278-95-5
          Dalapon                                        75-99-0
          Dibromoacetic Acid  (DBAA)                     631-64-1
          Dichloroacetic Acid (DCAA)                     79-43-6
          Monobromoacetic Acid (MBAA)                    79-08-3
          Monochl.oroacetic Acid (MCAA)                   79-11-8
          Tribromoacetic Acid (TBAA)                     75-96-7
          Trichloroacetic Acid (TCAA)                    76-03-9

     1.2  This method is applicable to the determination of the target
          analytes over the concentration ranges typically found in drinking
          water (1,2,4).  Experimentally determined method detection limits
          (MDLs) for the above listed analytes are provided in Table 2.
          Actual MDLs may vary according to the particular matrix analyzed and
          the specific instrumentation employed.  The haloacetic acids are
          observed ubiquitously in chlorinated drinking water supplies at
          concentrations ranging from <1 to >50 M9/L-

     1.3  This method is designed for analysts skilled in liquid-liquid
          extractions, derivatization procedures and the .use of GC and
          interpretation of gas chromatograms.  Each analyst must demonstrate
          the ability to generate acceptable results with this method using
          the procedure described in Section 9.3.

     1.4  When this method is used for the analyses of waters from unfamiliar
          sources, it is strongly recommended that analyte identifications be
          confirmed by GC using a dissimilar column or by GC/MS if
          concentrations are  sufficient.

2.   SUMMARY OF METHOD

     2.1  A 40-mL volume of sample is adjusted to pH <0.5 and extracted with
          4-mL of methyl-tert-butyl-ether (MTBE).  The haloacetic acids that
          have been partitioned into the organic phase are then converted to

                                    552.2-2

-------
          their methyl esters by the addition of acidic methanol followed by
          slight heating.  The acidic extract is neutralized by a back-
          extraction with a saturated solution of sodium bicarbonate and the
          target analytes are identified and measured by capillary column gas
          chromatography using an electron capture detector (GC/ECD).
          Analytes are quantitated using procedural standard calibration.

3.   DEFINITIONS

     3.1  INTERNAL STANDARD (IS) — A pure analyte(s) added to a sample,
          extract, or standard solution in known amount(s) and used to measure
          the relative responses of other method analytes and surrogates that
          are components of the same sample or solution.  The internal
          standard must be an analyte that is not a sample component.

     3.2  SURROGATE ANALYTE (SA) — A pure analyte(s), which is extremely
          unlikely to be found in any sample, and which is added to a sample
          aliquot in known amount(s).before extraction or other processing and
          is measured with the same procedures used to measure other sample
          components.  The purpose of the SA is to monitor method performance
          with each sample.  <

     3.3  LABORATORY DUPLICATES (LD1 AND LD2) — Two aliquots of the same
          sample designated  as such in the laboratory.  Each aliquot is
          extracted,  derivatized and analyzed separately with identical
          procedures.  Analyses of LD1 and LD2 indicate the precision
          associated with laboratory procedures,  but not with sample
          collection, preservation,  or storage procedures.

     3.4  FIELD DUPLICATES (FD1 AND FD2J — Two separate samples collected at
          the same time and  place under identical  circumstances and treated
          exactly the same throughout  field and laboratory procedures.
          Analyses of FD1 and FD2 give a measure of the precision associated
          with sample collection, preservation and storage,  as  well  as with
          laboratory procedures.

     3.5  LABORATORY  REAGENT BLANK (LRB) — An aliquot of reagent water or
          other blank matrix that are  treated exactly as a sample including
          exposure to all  glassware,  equipment,  solvents,  reagents,  internal
          standards,  and surrogates  that are used  with other samples.   The LRB
          is used to  determine if method analytes  or other interferences are
          present in  the laboratory  environment,  the reagents,  or the
          apparatus.

     3.6  FIELD REAGENT BLANK (FRB)  —  An  aliquot  of reagent water or other
          blank matrix that  is placed  in a  sample  container  in  the laboratory
          and treated as a sample in all  respects,  including shipment to the
          sampling site,  exposure to sampling site  conditions,  storage,
          preservation and all  analytical  procedures.   The purpose of the FRB
          is to determine  if method  analytes  or  other interferences  are
          present in  the field environment.


                                   552.2-3

-------
3.7  LABORATORY FORTIFIED BLANK (LFB) ~ An aliquot of reagent water or
     other blank matrix to which known quantities of the method analytes
     are added'in the laboratory.  The LFB is analyzed exactly like a
     sample, and its purpose is to determine whether,the methodology is
     in control, and whether the laboratory is capable of making accurate
     and precise measurements.

3.8  LABORATORY FORTIFIED SAMPLE MATRIX (LFM) — An aliquot of an
     environmental sample to which known quantities of the method
     analytes are added in the laboratory.  The LFM is analyzed exactly
     like -a sample, and its purpose is to determine whether the sample
     matrix contributes bias to the analytical results.  The background
     concentrations of the analytes in the sample matrix must be
     determined in a separate aliquot and the measured values in the LFM
     corrected for background concentrations.

3.9  STOCK STANDARD SOLUTION (SSS) — A concentrated solution containing
     one or more method analytes prepared in the laboratory using assayed
     reference materials or purchased from a reputable commercial source.

3.10 PRIMARY DILUTION STANDARD SOLUTION (PDS) — A solution of several
     analytes prepared in the laboratory from stock standard solutions
     and diluted as needed to prepare calibration solutions and other
     needed analyte solutions.

3.11 CALIBRATION STANDARD (CAL) — A solution prepared from the primary
     dilution standard solution and stock standard solutions of the
     internal standards and surrogate analytes.  The CAL solutions are
     used to calibrate the instrument response with respect to analyte
     concentration.

3.12 QUALITY CONTROL SAMPLE (QCS) — A solution of method analytes of
     known concentration which is used to fortify an aliquot of reagent
     water or sample matrix.  The QCS is obtained from a source external
     to the laboratory and different from the source of calibration
     standards. It is used to check laboratory performance with
     externally prepared test material si

3.13 LABORATORY PERFORMANCE CHECK SOLUTION (LPC) — A solution of
     selected method analytes used to evaluate the performance of the
     instrumental system with respect to a defined set of method
     criteria.

3.14 METHOD DETECTION LIMIT (MDL) — The minimum concentration of an
     analyte that can be identified, measured and reported with 99%
     confidence that the analyte concentration is greater than zero.

3.15 MATERIAL  SAFETY DATA SHEET  (MSDS) — Written information provided  by
     vendors concerning a chemical's toxicity, health hazards, physical
     properties,  fire and reactivity da;ta including storage, spill, and
     handling  precautions.
                               552.2-4

-------
     3.16 ESTIMATED  DETECTION  LIMIT  (EDL)  --  Defined  as  either the MDL or a
          level of a compound  in  a sample  yielding  a  peak  in the final extract
          with a  signal  to  noise  (S/N)  ratio  of  approximately 5, whichever is
          greater.

     3.17 PROCEDURAL STANDARD  CALIBRATION  —  A calibration method where
          aqueous calibration  standards are prepared  and processed (e.g.
          purged, extracted and/or derivatized)  in  exactly the same manner as
          a sample.   All steps  in the process from  addition of sampling
          preservatives  through instrumental  analyses are  included in the
          calibration.   Using  procedural standard calibration compensates for
          any inefficiencies in the processing procedure.

     3.18 CONTINUING  CALIBRATION CHECK  (CCC)  -- A calibration standard con-
          taining one or more method analytes, which  is analyzed periodically
          to verify  the  accuracy of the existing calibration curves or re-
          sponse factors for those analytes.

4.   INTERFERENCES
     4.1
 Method  interferences  may  be  caused  by  contaminants  in  solvents,
 reagents,  glassware and other.sample processing  apparatus  that  lead
 to  discrete  artifacts or  elevated baselines  in chromatograms.   All
 reagents  and  apparatus must  be  routinely demonstrated  to be  free
 from  interferences under  the conditions of the analysis by analyzing
 laboratory reagent blanks  as described in Section 9.5.  Subtracting
 blank values  from sample  results is not permitted.

 4.1.1   Glassware must be  scrupulously cleaned.   Clean  all  glassware
        as  soon as possible after use by thoroughly  rinsing with the
        last solvent used  in  it.  Follow by washing  with hot water
        and detergent  and thorough rinsing with tap  water and reagent
        water.  Drain  and heat in an oven or muffle  furnace at 400°C
        for 1  hr.  Do  not heat volumetric ware but instead  rinse
        three  times with HPLC grade or better acetone.  Thorough
        rinsing with reagent  grade acetone may be substituted for the
        heating provided method  blank analysis confirms no  background
        interferant contamination is present.   Thermally stable
        materials such as PCBs may not be eliminated by these treat-
        ments.  After  drying  and cooling,  store'glassware in a clean
        environment free of all  potential  contamination.  To prevent
        any accumulation of dust or other contaminants, store glass-
       ware inverted  or capped with aluminum foil.

4.1.2  The use of high purity reagents  and solvents helps to mini-
       mize interference problems.   Each  new bottle of solvent
       should be analyzed  before use.   An  interference free solvent
        is a solvent  containing no peaks yielding  data at > MDL
        (Table 2)  and at the retention  times  of the  analytes of
       interest.   Purification of solvents by distillation  in all-
       glass  systems may be required.
                                   552.2-5

-------
     4.2  Interfering contamination may occur when a sample containing low
          concentrations of analytes is analyzed immediately following a
          sample containing relatively high concentrations of analytes.
          Routine between-sample rinsing of the sample syringe and associated
          equipment with MTBE can minimize ;sample cross-contamination.  After-
          analysis of a sample containing high concentrations of analytes,  one
          or more injections of MTBE should be made to ensure that accurate
          values are obtained for the next sample.

     4.3  Matrix interferences may be caused by contaminants that are coex-
          tracted from the sample.  The extent of matrix interferences will
          vary considerably from source to ;source, depending upon the water
          sampled.  Analyte identifications should be confirmed using the
          confirmation column specified in Table 1 or by GC/MS if the concen-
          trations are sufficient.

     4.4  Bromochloroacetic acid coelutes with an interferant on the DB-1701
          confirmation column.  The interferant has been tentatively identi-
          fied as dimethyl sulfide.  However, because of the difference in
          peak shapes, the peak for the ester of BCAA tends to "ride on" the .
          interferant peak and quantitative confirmation can be performed by
          manual integration that includes only the peak area of the target
          ester.

     4.5  Methylation using acidic methanol results in a partial decarboxyl-
          ation of tribromoacetic acid (8).  Therefore a substantial peak for
        '  bromoform will be observed in the chromatograms.  Its elution does
          not, however, interfere with any other analytes.  Furthermore, this
          demonstrates the need for procedural standards to establish the
          calibration curve by which unknov/n samples are quantitated.
5.   SAFETY
     5.1  The toxicity or carcinogenicity of each reagent used in this method
          has not been precisely defined; however, each chemical compound must
          be treated as a potential health hazard.  From this viewpoint,
          exposure to these chemicals must be minimized.  The laboratory is
          responsible for maintaining a current awareness file of OSHA regula-
          tions regarding the safe handling of the chemicals specified in this
          method.  A reference file of material safety data sheets should also
          be made available to all personnel involved in the chemical analy-
          sis.  Additional references to laboratory safety are available and
          have been identified (9-11) for the information of the analyst.

     5.2  The toxicity of the extraction solvent, MTBE, has not been well
          defined.  Susceptible individuals may experience adverse affects
          upon skin contact or inhalation of vapors.  Therefore protective  •
          clothing and gloves should be used and MTBE should be used only in a
          chemical fume hood or glove box.  The same precaution applies to
          pure standard materials.
                                    552.2-6

-------
6.   APPARATUS AND EQUIPMENT

     6.1  SAMPLE CONTAINERS — Amber glass bottles, approximately 50 ml
       ,   fitted with Teflon-lined screw caps.                         '

     6.2  EXTRACTION VIALS— 60 ml clear glass vials with teflon-lined screw
          caps.

     6.3  VIALS  — Autosampler,  2.0 mL vials with screw or crimp cap and  a
          teflon-faced seal.         ,,   -.

     6.4  STANDARD SOLUTION STORAGE CONTAINERS - .-10-20-ml amber glass vials
          with teflon lined-screw caps.        .

     6.5  GRADUATED CONICAL CENTRIFUGE TUBES WITH TEFLON-LINED SCREW CAPS -
          15-mL  with  graduated 1  mL markings.

     6.6  BLOCK  HEATER (or SAND  BATH)  - Capable  of holding screw cap conical
          centrifuge  .tubes in  Section  6.4.

     6.7  PASTEUR  PIPETS  — Glass,  disposable.                       .

     6.8  PIPETS — 2.0 mL and 4.0  mL,  type  A,  TD,  glass.

     6.9-  VOLUMETRIC  FLASKS — 5  ml, 10 mL.

     6-10  MICRO SYRINGES  - 10 /tl_,  25  fil, 50 /iL,  100 0L, 250 fil,  500V and


     6.11  BALANCE —  analytical,  capable  of weighing to 0.0001 g.

     6.12  GAS  CHROMATOGRAPH — Analytical system complete  with gas chromato-
          graph equipped  for electron  capture detection, split/splitless
          capillary or direct  injection,  temperature programming, differential
          flow control, and with  all required accessories  including syringes
          analytical  columns,  gases and strip-chart recorder.  A data system'
          is recommended  for measuring  peak areas.  An autoinjector is recom-
        . mended  for  improved precision of analyses.  The gases flowing
         through the electron capture detector should be vented through the
         laboratory  fume hood system.

    6.13 PRIMARY GC COLUMN - DB-5.625 [fused silica capillary with chemical-
          ly bonded (5% phenyl)-methylpolysiloxane)] or equivalent bonded
         fused silica column,  30m x 0.25mm ID, 0.25 /im film thickness.  '

    6.14 CONFIRMATION GC COLUMN -- DB-1701  [fused silica capillary with
         chemically bonded (14% cyanopropylphenyl)-methylpolysiloxane)l or
         equivalent bonded, fused silica column,  30 m x 0.25 mm ID,  0.25  urn
         film thickness.
                                  552.2-7

-------
7.   REAGENTS AND STANDARDS

     7.1  REAGENT WATER — Reagent water is defined as a water in which an
          interference is not observed > to the MDL of each analyte of inter-
          est.

          7.1.1  A Millipore Super-Q water system or its equivalent may be
                 used to generate deionized reagent water.  Distilled water
                 that has been passed through granular charcoal may also be
                 suitable.

          7.1.2  Reagent water is monitored through analysis of the labora-
                 tory reagent blank (Section 9.5).

     7.2  SOLVENTS

          7.2.1  METHYL-TERT-BUTYL ETHER — High purity, demonstrated to be
                 free of analytes and interferences, redistilled in glass if
                 necessary.

          7.2.2  METHANOL — High purity, demonstrated to be free of
                 analytes and interferences.

          7.2.3  ACETONE — High purity, demonstrated to be free of analytes
                 and interferences.

     7.3  REAGENTS

          7.3.1  SODIUM SULFATE, Na2S04 —  (ACS)  granular*  anhydrous.   If
                 interferences are observedj.it may be necessary to heat the
                 sodium sulfate in a shallow tray at 400°C for up to 4 hr. to
                 remove phthalates and other interfering organic substances.
                 Alternatively, it can be extracted with methylene chloride in
                 a Soxhlet apparatus for 48 hr.  Store in a capped glass
                 bottle rather than a plastic container.

          7.3.2  COPPER II SULFATE PENTAHYDRATE, CuS04'5H20  --  ACS  re-
                 agent grade.

          7.3.3  SODIUM BICARBONATE, NaHC03 -- ACS reagent grade.

          7.3.4  AMMONIUM CHLORIDE, NH4C1 — ACS reagent grade, used to
                 convert  free chlorine to monochloramine. Although this
                 is not the traditional dechlorination mechanism,  ammoni-
                 um chloride  is categorized as a dechlorinating  agent  in
                 this method.

     7.4  SOLUTIONS

          7.4.1  10% H2S04/METHANOL SOLUTION -- Use caution when prepar-
                 ing sulfuric acid solutions.  To prepare  a 10%  solution,
                 add 5 mL sulfuric acid dropwise  (due to  heat  evolution)

                                    552.2-8

-------
     7.4.2
       to 20-30 mL methanol contained in a 50.0 mL volumetric
       flask that has been placed in a cooling bath.  Then
       dilute to the 50.0 ml mark with methanol.

       SATURATED SODIUM BICARBONATE SOLUTION — Add sodium
       bicarbonate to a volume of water, mixing periodically
       until the solution has reached saturation.
7.5  STANDARDS
7.5.1
            1,2,3-TRICHLOROPROPANE, 99+% — For use as the internal
            standard.  Prepare an internal standard stock standard solu-
            tion of 1,2,3-trichloropropane in MTBE at a concentration of
            approximately 1 mg/mL.  From this stock standard solution,
            prepare a primary dilution standard in MTBE at a concentra-
            tion of 25
     7.5.2  2,3-DIBROMOPROPIONIC ACID,  99+% — For use as a surrogate
            compound.   Prepare a surrogate stock standard solution of
            2,3-dibromopropionic acid in MTBE at a concentration of
            approximately 1 mg/mL.   From this stock standard solution,
            prepare a  primary dilution  standard in MTBE at a concentra-
            tion of 10 /ig/mL.

     7.5.3  STOCK STANDARD SOLUTION (SSS)  .

            Prepare separate stock  standard  solutions  for each  analyte  of
            interest at a concentration of 1-5 mg/mL in MTBE.   Method
            analytes may be obtained as neat materials or ampulized
            solutions  (> 99% purity) from  a  number of  commercial  suppli-
            ers.   These stock standard  solutions shcjld be stored at  -
            10°C  and protected from  light.   They  are stable  for at  least
            one  month  but should be checked  frequently for signs  of
            evaporation.

            7.5.3.1. For analytes which are  solids in  their pure  form,
                    prepare stock  standard  solutions  by accurately
                    weighing approximately  0.01  to  0.05 grams  of pure
                    material  in a  10.0 mL volumetric  flask.  Dilute  to
                    volume with MTBE.   When a compound purity  is assayed
                    to be 96% or greater, the weight  can  be used without
                    correction  to  calculate the  concentration  of the
                    stock standard. .

            7.5.3.2. Stock standard  solutions  for analytes  which  are
                    liquid in their pure form at room temperature  can
                    be accurately  prepared  in the following manner.

            7.5.3.3. Place about 9.8 mL  of MTBE  into a 10.0  mL volumetric
                    flask.   Allow the  flask to  standj  unstoppered, for
                    about 10 minutes to allow solvent film  to evaporate
                              552.2-9

-------
                from the inner walls of the volumetric,  and weigh to
                the nearest 0.1 mg.

       7.5.3.4. Use a 10 /iL syringe and immediately add  10.0 /iL of
                standard material to; the flask by keeping the
                syringe needle just above the surface of the MTBE.
                Be sure that the standard material  falls dropwise
                directly into the MT;BE without contacting the inner
                wall of the volumetric.

       7.5.3.5. Reweigh, dilute to v'olume, stopper, then mix by
                inverting the flask several times.   Calculate the
                concentration in milligrams' per milliliter from the
                net gain in weight.

7.5.4  PRIMARY DILUTION STANDARD (PDS) — Prepare the primary
       dilution standard solution by combining and diluting stock
       standard solutions with MTBE .(the surrogate' stock standard   ..
       solution was prepared in Section 7.5.2).  This primary
       dilution standard solution should be stored at -10°C and
       protected from light.  It is stable for at least one month
       but should be checked before :use for signs of evaporation.
       As a guideline to the analyst, the primary dilution standard
       solution used in the validation of this method is described
       below.

                                  Concentration.
       Monochloroacetic acid                60
       Monobromoacetic acid                 40
       Dalapon                              40
       Dichloroacetic acid                  60
       Trichloroacetic acid                ,20
       Bromochloroacetic acid               40
       Dibromoacetic acid                   .20
       Bromodichloroacetic acid             40
       Chlorodibromoacetic acid             100
       Tribromoacetic acid                  200
       2,3-Dibromopropionic acid  (surr.)    100

       This primary dilution standard  is used to prepare calibration
       standards, which comprise  five  concentration levels of each
       analyte with the lowest standard being at or near the MDL of
       each analyte.  The concentrations of the other  standards
       should define a range containing the expected sample
       concentrations or the working range  of the detector.
       NOTE:   When  purchasing  commercially  prepared  standards,  solu-
       tions  prepared  in  methanol  must  not  be  used because  it  has
       been found that the  haloacetic acids are  subject  to
       spontaneous  methylation when  stored  in  this solvent  (12).

                          552.2-10

-------
          7.5.6
 Furthermore, tribromoacetic acid has been found to be unsta-
 ble in methanol because it undergoes decarboxylation when
 stored in this solvent.

 7.5.4.1.  Include the surrogate analyte,  2,3-dibromopropionic
          acid, within the primary dilution standard prepared
          in Section 7.5.4.  By incorporating the surrogate
          into the primary dilution standard, it is diluted
          alongside the target analytes in the standard
          calibration curve.  This is done so that the peaks
          for the surrogate and the ester of chlorodibromo-
          acetic acid, which elute fairly closely,  are
          relatively close in size and adequate resolution is
          therefore insured.  Furthermore,  if a sample should
          have a very large concentration of chlorodibromo-
          acetic acid, it may be impossible to obtain an
          accurate measurement of surrogate recovery.   If this
          happens,  reextraction with a higher surrogate
          concentration would be an option.

 LABORATORY  PERFORMANCE CHECK STANDARD (LPC)  — A low level   '
 calibration standard can. serve as the LPC  standard.
8.   SAMPLE COLLECTION. PRESERVATION AND STORAGE

     8.1  SAMPLE VIAL PREPARATION
          8.1.1
          8.1.2
Grab  samples must  be  collected  in  accordance with conven-
tional  sampling  practices  (13)  using  amber glass containers
with  TFE-lined screw-caps  and capacities of at least 50 ml.

Prior to shipment  to  the field, add crystalline or granular
ammonium chloride  (NH4C1) to the sample container in an
amount  to produce  a concentration  of  100 mg/L in the sample.
For a typical 50 mL sample, 5 mg of ammonium chloride  is
added.

NOTE:   Enough ammonium chloride must  be added to the sample
to convert the free chlorine residual in the sample matrix to
combined chlorine. Typically, the  ammonium chloride
concentration here will accomplish that.  If high doses of
chlorine are used, additional ammonium chloride may be re-
quired.
     8.2   SAMPLE COLLECTION
          8.2.1
          8.2.2
Fill sample bottles to just overflowing but take care not to
flush out the ammonium chloride.

When sampling from a water tap, open the tap and allow the
system to flush until the water temperature has stabilized
(usually about 3-5 minutes).  Remove the aerator so that no

                  552.2-11

-------
                 air bubbles can be visibly detected  and  collect  samples  from
                 the flowing system.

          8.2.3  When sampling from an open body of water,  fill  a 1-quart
                 wide-mouth bottle or 1-liter beaker  with sample  from a
                 representative area, and carefully fill  sample  vials from the
                 container.

          8.2.4  After collecting the sample in the bottle containing the
                 ammonium chloride, seal  the bottle and agitate  by hand for 1
                 min.

     8.3  SAMPLE STORAGE/HOLDING TIMES

          8.3.1  Samples must be iced or refrigerated at  4°C  and  maintained at
                 these conditions away from light until extraction.  Synthetic
                 ice (i.e., blue ice) is not recommended.  Holding studies
                 performed to date have suggested that, in samples preserved
                 with NH4C1, the analytes are stable  for  up to 14 days.   Since
                 stability may be matrix dependent, the analyst  should verify
                 that the prescribed preservation technique is  suitable for
                 the samples under study.

          8.3.2  Extracts  (Section 11.2.7) must be stored at 4°C  or less  away
                 from light in glass vials with Teflon-lined caps.  Extracts
                 must be analyzed within 7 days from  extraction  if stored at
                 4°C or within 14 days if stored at -10°C or less.

9.   QUALITY CONTROL

     9.1  Each laboratory that uses  this method is required to operate a
          formal quality control  (QC) program.  Minimum quality control
          requirements are  monitoring the laboratory performance check stan-
          dard, initial demonstration of laboratory capability,  performance of
          the method detection limit study,  analysis  of laboratory reagent
          blanks and laboratory fortified sample matrices,  determination of
          surrogate compound recoveries  in ;each sample and blank, monitoring
          internal standard peak  area or height in each sample,  blank and CCC,
          and analysis of  QC samples.  Additional QC practices may be added.

     9.2  LABORATORY PERFORMANCE  CHECK STANDARD (LPC)

          At the beginning  of  an  analysis set, prior to any calibration
          standard or sample analysis and after an initial solvent analysis, a
          laboratory performance  check standard must be analyzed.  This  check
          standard insures  proper  performance  of the GC by evaluation of the
          instrument parameters of detector  sensitivity, peak symmetry,  and
          peak resolution.  It  furthermore ;serves  as a check on the  continuity
          of the instrument's  performance.   In regards to  sensitivity, it
          allows the analyst to ascertain that this parameter has  not changed
          drastically since the analysis of  the MDL study.  Inability to
          demonstrate acceptable  instrument  performance  indicates  the need  for

                                   552.2-12

-------
                                                                                      I
     re-evaluation of the instrument system.  Criteria are listed in
     Table 8.

     9.2.1  The sensitivity requirement is based on the EDLs published in
            this method.  If laboratory EDLs differ from those listed in
            Table 2, concentrations of the LPC standard may be adjusted
            to be compatible with the laboratory EDLs.

     9.2.2  If column or chromatographic performance cannot be met, one
            or more of the following remedial actions should be taken.
            Break off approximately 1 meter of the injector end of the
            column and re-install, install a new column,  adjust column
            flows or modify the oven temperature program.

9.3  INITIAL DEMONSTRATION OF CAPABILITY (IDC)

     9.3.1  Calibrate for each analyte of interest as specified in
            Section 10.   Select a representative fortification
            concentration for each of the target analytes.
            Concentrations near those in Table 4 are recommended.
            Prepare 4-7  replicates laboratory fortified blanks by adding
            an appropriate aliquot of the primary dilution  standard or
            quality control  sample to reagent water.   (This reagent water
            should  contain ammonium chloride at the same  concentration
            as that specified for samples as per Section  8.1.2.)   Analyze
            the LFBs according to the method beginning in  Section 11.

     9.3.2  Calculate the mean percent recovery and the standard  devia-
            tion of the  recoveries.   For each analyte, the  mean recovery
            value,  expressed as a percentage of the true  value, must fall
            in the range of 80-120% and the relative standard deviation
            should be less than 20%.   For those compounds  that meet these
            criteria,  performance is  considered acceptable  and sample
            analysis may begin.  For  those compounds  that fail  these
            criteria,  this procedure  must be repeated using 4-7 fresh
            samples until  satisfactory performance has been demonstrated.
            Maintain these data on file to demonstrate initial
            capabilities.

     9.3.3  Furthermore,  before processing any samples, the analyst must
            analyze at least one laboratory reagent blank to demonstrate
            that all  glassware and reagent interferences are under
            control.

     9.3.4  The initial  demonstration  of capability is used primarily  to
            preclude a laboratory from analyzing  unknown samples  via a
            new,  unfamiliar  method prior to obtaining some  experience
            with it.   As  laboratory personnel  gain experience with this
            method,  the  quality of data should improve beyond those re-
            quired  here.
                              552.2-13

-------
     9.3.5  The analyst is permitted to modify GC columns,  GC conditions,
            internal standard or surrogate compounds.  Each time such
            method modifications are made, the analyst must repeat the
            procedures in Section 9.3.1 through Section 9.3.4 and Sect.
            9.4.

9.4  METHOD DETECTION LIMIT STUDY (MDL)

     9.4.1. Prior to the analysis of any: field samples, the method
            detection limits must be determined.  Initially, estimate the
            concentration of an analyte which would yield a peak equal to
            5 times the baseline noise and drift.  Prepare seven
            replicate laboratory fortified blanks at this estimated
            concentration with reagent water that contains ammonium
            chloride at the same concentration as that specified for
            samples as per Section 8.1.2'.  Analyze the LFB's according to
            the method beginning in Section 11.

     9 4.2. Calculate the mean recovery land the standard deviation for
            each  analyte.  Multiply the student's t  value at 99% confi-
            dence and n-1 degrees of freedom  (3.143  for seven replicates)
            by  this standard deviation to yield a statistical estimate of
            the detection limit.  This calculated value is the MDL.

     9.4.3. Since the statistical estimate is  based  on the preci-  sion
            of  the  analysis, an  additional estimate  of detection can  be
            determined based upon the  noiise and drift  of the baseline as
            well  as precision.   This estimate  is the EDL (Table  2).

9.5  LABORATORY REAGENT  BLANKS  (LRB) —   Each  time a set of  samples  is
     extracted  or reagents  are  changed,  a LRB  must be  analyzed.   If  the
     LRB  produces an  interferant peak  within  the retention  time  window
     (Section  12.3) of any  analyte  that  would  prevent  the  determination
     of that analyte  or  a peak  of concentration  greater than the MDL for
     that analyte,  the analyst  must determine  the  source of contamination
     and  eliminate  the interference before processing  samples.   Field
     samples of an  extraction  set associated  with  an LRB that  has  failed
     the  specified  criteria are considered suspect.

     NOTE:   Reagent water containing ammonium chloride at  the  same
     concentrations as  in the  samples  (Section 8.1.2)  is used  to prepare
     the  LRB.

 9.6  LABORATORY FORTIFIED BLANK (LFB)  — Since this  method utilizes
     procedural calibration standards, which  are fortified reagent water,
     there is  no  difference between the LFB  and the  continuing
     calibration  check standard.  Consequently,  the  analysis of an LFB is
     not  required (Section  10.2).
                               552.2-14

-------
 9.7.2.
 9.7  LABORATORY FORTIFIED SAMPLE MATRIX  (LFM)

      9.7.1  Chlorinated water .supplies will usually contain significant
             background concentrations of several method analytes  espe-
             /T?«l? dichloroacetic ^id (DCAA) and trichloroacetic acid
             (ICAA).  The concentrations of these acids may be equal to or
             greater than the fortified concentrations.  Relatively poor
             accuracy and precision may be anticipated when a large
             background.must be subtracted.   For many samples,  the concen-
             trations may be so high that fortification may lead to a
             final  extract with instrumental responses exceeding the
             linear range of the electron capture detector.  If this
             occurs, the extract must be diluted.  In spite of these
             problems,  sample sources should be fortified and analyzed as
             described  below.  By fortifying sample matrices and calcu-
             lating analyte recoveries,  any  matrix induced analyte bias is
             evaluated.

             The laboratory must add known concentrations of analytes to
             one sample  per extract!on set or a minimum of 10%  of the
             samples, whichever is  greater.   The concentrations should be
             equal  to or  greater than the  background  concentrations in the
             sample selected  for fortification.   If the fortification
             level  is less  than the  background  concentration, recoveries
             are not reported.   Over time, samples  from all  routine sample
             sources should be  fortified.

             Calculate the  mean  percent  recovery,  R,  of the  concentration
             for each analyte,  after  correcting  the total mean  measured
             concentration, A,  from  the  fortified  sample  for the  back-
             ground  concentration, B,  measured  in.the  unfortified  sample
             i.e.:                                                    K

                     R = 100 (A - B)  /  C,

             where C is the fortifying concentration.   In order for the
             recoveries to  be considered acceptable, they must fall
             between 70% and  130% for  all the target analytes.

             If  a recovery .falls outside of this acceptance  range, a
            matrix  induced bias can be assumed for the respective analyte
             and the data for that analyte must be reported  to the data
            user as suspect due to matrix effects.

9.8  ASSESSING SURROGATE RECOVERY

     The surrogate analyte is fortified into the aqueous portion of all
     continuing calibration standards, samples and laboratory reagent
     blanks.   The surrogate is  a means of assessing method performance in
     every analysis from extraction to final chromatographic performance
9.7.3
9.7.4
                         552.2-15

-------
     9.8.1  When surrogate recovery from a sample,  blank or CCC is < 70%
            or > 130%, check (1) calculations to locate possible errors,
            (2) standard solutions for degradation,  (3) contamination,
            and (4) instrument performance.   If those steps do not reveal
            the cause of the problem, Reanalyze the extract.

     9.8.2  If the extract reanalysis meets  the surrogate recovery
            criterion, report only data for  the reanalyzed extract.

     9.8.3  If the extract reanalysis fails  the 70-130% recovery
            criterion, the analyst should check the calibration by
            analyzing the most recently acceptable continuing calibration
            check standard.  If the CQC fails the criteria of Section
            10.2.1, recalibration is In order per Section 10.1.  If the
            CCC is acceptable, it may be necessary to extract another
            aliquot of sample.  If the sample re-extract also fails the
            recovery criterion, report all data for that sample as
            suspect.

9.9  ASSESSING THE INTERNAL STANDARD

     9.9.1. The analyst must to monitor the  IS response (peak area or
            peak height) of all injections during each analysis day.
            A mean IS response should be determined from the five point
            calibration curve.  The IS response for any run should not
            deviate from this mean IS response by more than 30%.  It is
            also acceptable if the IS response of a injection is within
            15% of the daily continuing calibration standard IS response.

     9.9.2  If a deviation greater than this occurs with an individual
            extract, optimize instrument performance and inject a second
            aliquot of that extract.

            9.9.2.1   If the reinjected aliquot produces an acceptable
                      internal standard response, report results for that
                      aliquot.

            9.9.2.2   If a deviation of greater than 30% is obtained for
                     the reinjected extract, the analyst should check the
                      calibration by analyzing the most recently
                      acceptable CCC.  ;If the CCC fails the criteria of
                      Section 10.2.1, recalibration is  in order per
                      Section 10.1.   If the CCC is acceptable, analysis  of
                      the sample should be repeated beginning with Section
                      11, provided the sample is still  available.  Oth-
                      erwise, report results obtained from the reinjected
                      extract, but annotate as suspect.

9.10 QUALITY CONTROL  SAMPLE  (QCS) — At least quarterly, analyze a QCS
     from an external source.   If measured analyte concentrations are not
     of acceptable accuracy, check the entire analytical procedure to
     locate and correct the problem  source.

                              552.2-16

-------
     9.11 The laboratory may adapt additional QC practices for use with this
          ?ho ±H   f tteC,fiC P:actices that are most productive depend upon
          the needs of the laboratory and the nature of the samples   For
          example, field or laboratory duplicates may be analyzed to assess
          the precision of the environmental measurements or field reaqent
          blanks may be used to assess contamination of samples under site
          conditions, transportation and storage.                   '»•«*_

10.   CALIBRATION AND STANDARDIZATION

     10.1 INITIAL CALIBRATION CURVE

          10.1.1 Calibration is performed by extracting procedural  standards,
                 i.e ;  fortified reagent water,  by the procedure  set forth in
                 Section  11.   A five-point calibration curve  is to  be prepared
                 by diluting the primary dilution standard  into MTBE at  the
                 appropriate levels.   The desired amount of each  MTBE
                 calibration standard  is added to separate  40  mL  aliquots  of
                 reagent  water  to  produce a  calibration  curve  ranging from the
                 detection  limit to approximately 50  times  the detection
                 limit.   (These MTBE calibration  standards  should be  prepared
                 so that  20 /jl  or  less of the solution  is added the water
                 aliquots.)  Also,  the reagent water  used for  the procedural
                 standards  contains ammonium chloride  at the same concentra-
                 tion as  that in the samples as per Section 8.1.2.

          10.1.2  Establish  GC operating  parameters equivalent  to the  suggested
                 specifications  in  Table  1.  The  GC system must be calibrated
                 using the  internal standard (IS) technique.   Other columns or
                 conditions may be  used  if equivalent or better performance
                 can be. demonstrated.

          10.1.2  Five calibration standards  are required. The lowest should
                 contain the analytes at a concentration near to but greater
                 than the MDL (Table 2) for each compound. The others should
                 be evenly distributed throughout the concentration ranae
                expected in the samples.

         10.1.3  Inject  2 >L of each calibration  standard extract  and tabulate
                peak height or area response and concentration for each
                analyte and the internal standard.

         10.1.4 Generate a  calibration curve by  plotting the  area ratios
                (V^is) against the concentration Ca  of the five  calibration
                standards where

                         Aa is.the  peak  area of the analyte.
                         A}s is the peak area of the internal  standard.
                         Ca is  the  concentration  of the analyte.
                                  552.2-17

-------
           This curve  can  be defined  as either first or second order.
           Also, the working calibration curve must be verified daily by
           measurement of  one  or more calibration  standards  (Section
           10.2).   If  the  response  for any  analyte falls  outside the
           predicted response  by more than  30%,  the calibration check
           must be  repeated using a freshly prepared calibration stan-
           dard.  Should the retest fail, a new  calibration  curve  must
           be  generated.

     10.1.5 Alternately, an average  relative response factor  can be
           calculated  and  used for  quantitation.   Relative response
           factors  are calculated for each  analyte at  the five
           concentration levels using the equation below:
            RRF
(Aa)(Cis)

(Ais)(Ca)
            If the RRF value over the working range is constant (<20%
            RSD),  the RRF can be assumed to be invariant and the average
            RRF used for calculations.  Also, the average RRF must be
            verified daily by measurement of one or more calibration
            standards (Secti-on 10.2).  If the RRF for the continuing
            calibration standard deviates from the average RRF by more
            than 30%, the calibration check must be repeated using a
            freshly prepared calibration standard.  Should the retest
            fail,  a new calibration curve must be generated.

     10.1.6 A data system may be used to collect the chromatographic
            data,  calculate relative response factors, or calculate
            linear or second order calibration curves.

10.2 CONTINUING CALIBRATION CHECK (CCC)

     10.2.1 At least one CCC must be extracted with each set of samples.
            A CCC must be analyzed at the beginning of each, analysis set,
            after every tenth sample analysis and after the final sample
            analysis, to ensure that the instrument is still within
            calibration.  These checks should be  at two different
            'concentration levels.  Calculate analyte recoveries for all
            target analytes.  In order for the calibration check to be
            considered valid and subsequently for the preceding ten
            samples  to be considered  acceptable with respect to
            calibration, recoveries must fall between 70% and  130% for
            all the  target  analytes.

            NOTE: Continuing calibration check standards need  not
            necessarily  be  different  extracts but can be injections from
            the same extract as long  as  the  holding time requirements
            (Sect. 8.3.2) are met.
                               552.2-18

-------
                                                                                          I
          10.2.2  If this^.critena cannot be met, the continuing calibration
                  check standard extract is re-injected in order to determine
                  if the response deviations observed from the initial analysis
                  are repeated.  If this criteria still cannot be met, a second
                  CCC should be extracted and analyzed or a CCC that has
                  already been analyzed and has been found to be acceptable
                  should be run.  If this second CCC fails, then the instrument
                  is considered out of calibration and needs to be
                  recalibrated.

11.   PROCEDURE

     11.1 SAMPLE EXTRACTION

          11.1.1 Remove the samples  from storage (Sect.  8.3.1)  and allow them
                 to equilibrate to room temperature.

          11.1.2 Place  40  ml  of the  water  sample into  a  precleaned 60  ml glass
                 vial with a  teflon-lined  screw cap  using  a  graduated
                 cylinder.

          11.1.3 Add  20 ML of  surrogate standard  (10.0 ,/ig/mL  2,3-dibromo-
                 propionic acid  in MTBE per Section  7.5.2).

                 NOTE:  When fortifying an aqueous sample with either
                 surrogate  or  target analytes contained in MTBE, be sure that
                 the needle of the syringe is well below the  level of  the
                 water.  After injection, cap the sample and  invert once
                 This insures  that the  standard solution is mixed well with
                 the water.

         11.1.4 Adjust the pH to less  than 0.5 by adding at  least 2 mL of
                concentrated  sulfuric  acid.   Cap, shake and then check the PH
                with a pH meter or narrow range pH paper.

         11.1.5 Quickly add approximately 2 g of copper II sulfate
                pentahydrate and shake until  dissolved.   This colors the
                aqueous phase blue and therefore allows  for the analyst to
                better distinguish between the aqueous phase and the organic
                phase in this micro  extraction.

         11.1.6 Quickly add 16 g of  muffled  sodium sulfate and  shake for 3 to
                5 minutes  until  almost all  is  dissolved.   Sodium sulfate is
                added  to  increase  the  ionic  strength of  the  aqueous  phase and
                thus  further  drive the haloacetic acids  into  the organic
                Pu S?:  uThe addition of this  salt and  the  copper II  sulfate
                should  be  done quickly so  that  the heat generated  from the
                addition of the  acid  (Section  11.1.4) will help  dissolve the
                S 3 I t S .
                                  552.2-19

-------
     11 1  7 Add 4.0 ml MTBE and place  on  the  mechanical  shaker  for 30
       '  '   minutes.  (If hand-shaken,  two  minutes  is  sufficient  if
            performed vigorously).

     11.1.8 Allow the phases to separate  for  approximately 5 minutes.

11.2 METHYLATION

     11 2 1 Usinq a pasteur pipet,  transfer approximately 3 ml of the
            uppe? MTBE layer to a 15 ml graduated conical centrifuge
            tube.

     11.2.2 Add l.mL  10% sulfuric acid' in methanol  to each centrifuge
            tube.
n-2-3

       tained.
            snugly  into the heating 'block to ensure proper heat transfer.
            At this  stage, methyl ation of the method analytes is at-
            tained.

      11  2  4 Remove  the  centrifuge tubes  from the  heating block  (or sand
            bath)  and allow them to  cool before removing the caps.

      11  2  5 Add  4  ml saturated sodium bicarbonate solution to each
      11     centrifuge  tube  in 1 ml  increments    Exercise caution when
            adding the  solution because  the evolution  of C02 in this
            neutralization reaction  is rather  rapid.

      11  2  6  Shake  each  centrifuge tube for 2 minutes.   As the neutral -
             ization reaction moves to completion, it  is important  to
             continue to exercise caution by venting frequently  to  release
             the evolved C02.

      11  2 7 Transfer exactly 1.0 ml  of the upper MTBE layer to  an  auto-
             sampler vial.  A duplicate vial should be filled using the
             excess extract.

      11 2 8 Add 10 uL of  internal standard to the vial to be analyzed.
             (25 Mg/mL 1,2,3-trichloropropane in MTBE per Section 7.5.1).
      11 2 9 Analyze the  samples  as soon as possible.  The sample extract
             may  be stored  up  to  7 days if kept at 4°C or less or up to 14
             days if kept at -10°C or  less.  Keep the extracts away from
             light in  amber glass vials with Teflon-lined caps.

  11.3 GAS CHROMATOGRAPHY

      11 3.1 Table 1 summarizes  recommended GC  operating c°nditio"s.^c
             retention times  observed  using this method.  Figure  1  illus-
             trates the performance  of the  recommended  primary column  with
             the method analytes.  Figure  2  illustrates  the  performance of
                                552.2-20

-------
                  the recommended confirmation column with the method analytes
                  Concentrations of the analytes of these chromatograms are
                  those listed in Table 4 for the fortified reagent water
                  samples.  Other GC columns or chromatographic conditions may
                  be used if the requirements of Section 9 are met.

           11.3.2 Calibrate the system (Section 10.1) or verify the existing
                  calibration by analysis of a CCC daily as described in
                  Section 10.2.

           11.3.3 Inject 2 /il_ of the sample extract.   Record the resulting peak
                  sizes in area or height units.

           11.3.4 If the response for the peak exceeds the working range of the
                  system,  dilute the extract,  add an  appropriate additional
                  amount of .internal  standard  and reanalyze.   The analyst must
                  not extrapolate beyond  the calibration range established.

 12.   DATA  ANALYSIS  AND CALCULATIONS

      12.1  Identify  sample components  by  comparison of retention times  to
           retention data  from  the  calibration  standard  analysis.   If the
           retention time  of an unknown peak corresponds,  within limits  (Sec-
           tion  12.2),  to  the retention time of a  standard compound,  then the
           identification  is considered positive.  Calculate  analyte  concentra-
           tions  in  the  samples and  reagent  blanks from  the calibration  curves
           generated in  Section 10.1.                  '

      12.2  If an  average relative response factor  has  been calculated (Sect
           10.1.5),  analyte  concentrations in the  samples  and  reagent blanks
           are calculated  using the  following equation:


           c  - -(-A-a-Cis)
               (Ais)(RRF)

      12.3 The width of  the  retention time window used to make identifications
           should be based upon measurements of actual retention time varia-
          tions of  standards over the course of a day.  Three times the
          standard deviation of a retention time can be used to calculate a
          suggested window  size for a compound.  However, the experience of
          the analyst should weigh heavily  in the interpretation of
          chromatogram.

13.   METHOD PERFORMANCE

     13.1 In a single laboratory, recovery and precision data were obtained at
          three concentrations  in reagent water (Tables 3 and 4).  The  MDL and
          EDL data are given in Table 2.   In addition, recovery and precision
          data were  obtained at a medium  concentration for dechlorinated tap
          water (Table 5), high ionic strength reagent water (Table 6)  and
          high  humectant ground water (Table 7).

                                   552.2-21

-------
14.  POLLUTION PREVENTION

     14.1 This method utilizes a micro-extraction procedure which requires the
          use of very small quantities of organic solvents. This feature
          reduces the hazards involved with;the use of large volumes of poten-
          tially harmful organic solvents needed for conventional liquid-
          liquid extractions.  This method also uses acidic methanol as the
          derivatizing reagent.

     14.2 For information about pollution prevention that may be applicable to
          laboratory operations consult "Less is Better:  Laboratory Chemical
          Management for Waste Reduction" available from the American Chemical
          Society's Department of Government Relations and Science Policy,
          1155 16th Street N.W., Washington, D.C. 20036.

15.  WASTE MANAGEMENT

     15.1 Due to the nature of this method there is little need  for waste
          management.  No large volumes of solvents or hazardous chemicals are
          used.  The matrices of concern are finished drinking water or source
          water.  However, the Agency requires that laboratory waste manage-
          ment practices be conducted consistent with all applicable rules and
          regulations, and that laboratories protect the air, water, and land
          by minimizing and controlling all releases from fume hoods and bench
          operations.  Also compliance is required with any sewage discharge
          permits and regulations, particularly the hazardous waste identifi-
          cation rules and land disposal restrictions.  For further informa-
          tion on waste management, consult;"The Waste Management Manual for
          Laboratory Personnel" also available from the American Chemical
          Society at the address in Sect. 14.2.
16.
REFERENCES

1.
     4.
Quimby, B.D., Delaney, M.F., Uden. P.C. and Barnes, R.M.  Anal.
Chem. 52, 1980, pp. 259-263.

Uden, P.C. and Miller, J.W., J. Am. Water Works Assoc. 75, 1983, pp.
524-527.

Hodgeson, J.W. and Cohen, A.L. anjd Collins, J.D., "Analytical
Methods for Measuring Organic Chlorination Byproducts", Proceedings
Wate.r Quality Technology Conference (WQTC-16), St. Louis, MO, Nov.
13-17, 1988, American Water Works Association, Denver, CO, pp. 981-
1001.

Fair. P.S., Barth, R.C., "Comparison of the Microextraction
Procedure and Method 552 for the Analysis of HAAs and
Chlorophenols", Journal AWWA, November, 1992, pp. 94-98.
                                    552.2-22

-------
 7.
 8.
 9.
 10.
 11
12.
13.
14.
15.
 mc/u  iand  KT!3en'  S'."  A  SimP]ified  Technique  for  the  Measure-
 ment of Halogenated  Organic  Acids  in  Drinking  Water by  Electron
 Capture Gas Chromatography". Presented  at the  28th  Pacific  Confer-
 ence on Chemistry  and Spectroscopy, Pasadena,  CA, October,  1989

 Hodgeson,  J  W., Collins, J. D., and  Becker, D. A.,  "Advanced Tech-
 niques for the Measurement of Acidic  Herbicides and  Disinfection
 Byproducts in Aqueous Samples," Proceedings of the  14th Annual EPA
 Conference on Analysis  of Pollutants  in the Environment, Norfolk,
 VA., May 8-9, 1991.  Office of Water  Publication No. 821-R-92-001
 U.S. Environmental Protection Agency, Washington, DC, pp 165-194.

 Shorney,  Holly L. and Randtke,  Stephen J., "Improved Methods for
 Haloacetic Acid Analysis", Proceedings Water Quality Technology
 Conference, San Francisco, CA,  November 6-10,  1994,  American Water
 Works Association, pp 453-475.

 Peters  Rund J.B., Erkelens,  Corrie, De Leer,  Ed W.B. and De Galan
 Leo,  The Analysis of Halogenated Acetic Acids in Dutch Drinkina
 Water",   Wat.  Res., Vol.25,  No.4,  1991,  Great Britain,  pp 473-477.

 "Carcinogens-Working with Carcinogens",  Publication  No   77-206
 Department of Health, Education,  and Welfare,  Public Health Service
 Center for Disease Control,  National Institute of Occupational
 Safety and Health,  Atlanta,  Georgia, August 1977.

 "OSHA Safety  and  Health  Standards,  General Industry", (29CFR1910)
 OSHA 2206,  Occupational  Safety  and  Health  Administration,  Washing-
 ton,  D.C.   Revised January 1976.

 "Safety In  Academic Chemistry Laboratories",  3rd  Edition,  American
 Chemical Society  Publication, Committee  on Chemical  Safety,  Washing-
 ton,  U.L.,  1979.

 Xie   Yuefeng, Reckhow, David  A., and Rajan, R.V.,  "Spontaneous
 Methylation of  Haloacetic  Acids in  Methanolic Stock  Solutions"
 Environ. Sci. Techno!.,  Vol.27, No.6,  1993, pp!232-1234.

 ASTM Annual Book of Standards,  Part  31,  D3370,  "Standard Practice
 for Sampling Water",  American Society  for  Testing and Materials
 Philadelphia, PA, p.  76, 1980.         ,                        '

ASTM Annual Book of Standards, Part  31, D3694,  "Standard Practice
 for Preparation of Sample Containers and for Preservation"  American
Society for Testing and  Materials, Philadelphia, PA,  p.  679, 1980.

Glaser  J.  A., Foerst, D. L.,  McKee, G. D., Quave, S. A. and Budde,
W.  L., Environ.  Sci. Technol.  15,  1981, pp. 1426-1435.
                              552.2-23

-------
                    wai
                 •nuns
       in
       CM
       to

       in

       CO
       a
                                             WSCKT
       a:
       a.

       UJ
                                                        yvaa
                                                                                     S£ -
                                                                                     OE  -
        O

        co
        UJ
                        woa
                                                       W3i
Q


Z
a
HH
        I
        S
                                                                                     OZ
                                                      NOdVTva •
                     woa
                               •  WSH
>

Q


-------
                                         wai-
           •nuns
                               W803
 T—l




 00
                                                                           9£
 o
 o
o
I—I

&
                                                    waa
                                                     W3Q8
                                                                           oc
o
o
                                                                           sz
                           woa
ts>
UJ

-------
           TABLE 1.  RETENTION DATA AND CHROMATOGRAPHIC CONDITIONS
Analyte
Monochloroacetic Acid (MCAA)
Monobromoacetic Acid (MBAA)
Dichloroacetic Acid (DCAA)
Dalapon
Trichloroacetic Acid (TCAA)
Bromochloroacetic Acid (BCAA)
1,2,3-Trichloropropane (L.S.)
Dibromoacetic Acid (DBAA)
Bromodichloroacetic acid (BDCAA)
Chlorodibromoacetic acid (CDBAA)
2,3-Dibromopropionic acid (SURR)
Tribromoacetic Acid (TBAA)
Retention Time.
l
Column A
13103
17,15
. 17,80
19.08
22.67
23.15
23,70
31.38
32,18
41.57
41.77
49.22
min.
Column B
13.70
17.33
17.88
17.73
20.73
22.87
22.35
30.27
28.55
38.78
39.72
,47.08
Column A:  DB-5.625, 30 m x 0.25 mm i.d., 0.25 im film thickness,
           Injector Temp. = 200 C,  Detector Temp. = 260 C,  Helium
           Linear Velocity = 24 cm/sec at 35°C,  Splitless  injection
           with 30 s delay

Program:   Hold at 35°C for 10 min, ramp to 75°C at 5C°/min. and hold
           15 min., ramp to 100°C  at.  5C°/min. and hold 5 min,  ramp to
           135°C at 5C°/min. and hold 2 min.

Column B:  DB-1701, 30 m x 0.25 mm i.d., 0.25 /on film thickness,  Injec-
           tor Temp. = 200 C, Detector Temp. = 260°C, Linear Helium
           Velocity = 25 cm/sec at 35°C, splitless  injection with 30 s
           delay.

Program:   Hold at 35°C for 10 min, ramp to 75°C at 5C°/min. and hold
           15 min., ramp to  100°C  at  5C°/min. and  hold  5 min,  ramp to
           135°C at 5C°/min. and hold '0 min.
                                552.2-26

-------
                     TABLE 2.  ANALYTE ACCURACY AND PRECISION DATA
                                AND METHOD DETECTION LIMITS3
                                LEVEL  1  IN  REAGENT  WATER
Analyte
MCAA
MBAA
DC.AA
Dalapon
TCAA
BCAA
DBAA
BDCAA
CDBAA
TBAA
Fortified
Cone.
M9/L
0.600
0.400
0.600
0.400
0.200
0.400
0.200
0.400
1.00
2.00
Mean
Meas.
Cone.
M9/L
0.516
0.527
0.494
0.455
0.219
0.498
0.238
0.357
1.19 .
1.91
Std.
Dev.
M9/L
0.087
0.065
0.077
0.038
0.025
0.080
0.021
0.029
0.149
0.261
Rel.
Std.
Dev., %
17
12
16
8.4
11
16
8.8
8.1
12
14
Method
Detection
Limitb
M9/L
0.273
0.204
0.242
0.119
0.079
0.251
0.066
0.091
0.468
0.820
Estimated
Detection
Limit0
M9/L
0.60
0.20
0.24
0.40
0.20
0.25
0.20
0.40
0.75
1.5
a Produced  by  analysis  of seven  aliquots  of fortified  reagent  water.

b The MDL is a statistical estimate of the detection limit.   To determine the MDL for
each analyte,  the standard deviation of the mean concentration of the seven replicates
is calculated.  This standard deviation is then multiplied by the student's t-value at
99% confidence and n-1 degrees of freedom  (3.143 for seven replicates).  The result is
the MDL.

  The EDL is defined as either the MDL or a level  of a compound in a sample yielding a
peak in the final extract with a signal to noise (S/N) ratio of approximately 5,
whichever is greater.
                                       552.2-27

-------
                   TABLE 3.   ANALYTE ACCURACY AND PRECISION DATA8
                               LEVEL  2  IN  REAGENT  WATER
Fortified
Cone.
Analyte M9/L
Monochloroacetic Acid
Monobromoacetic Acid
Dichloroacetic Acid
Dalapon
Trichloroacetic Acid
Bromochloroacetic Acid
Dibromoacetic Acid
Bromodichloroacetic Acid
Chlorodibromoacetic Acid
Tribromoacetic Acid
1.50
1.00
1.50
1.00
0.500
1.00
0.500
1.00
2.50
5.00
Mean
Meas.
Cone.
M9/L
1.42
1.02
1.27
0.935
0.465
0.869
0.477
1.07
2.62
5.19
Std.
Dev.
M9/L.
: 0.103
0.051
, 0.122
0.087
0.048
0.049
0.044
0.098
0.150
0.587
Rel.
Std.
Dev., %
7.3
5.0
9.6
9.3
10
5.6
9.2
9.2
5.7
11 .
Mean
Recovery
%
94.7
102
84.7
93.5
93 .0
86.9
95.4
107
105
104
a Produced by the analysis of seven aliquots of fortified reagent water.
                                        552.2-28

-------
                  TABLE 4.   ANALYTE ACCURACY AND PRECISION DATA3
                              LEVEL 4 IN REAGENT WATER
Mean
Fortified Meas.
Cone. Cone.
Analyte /jg/L #g/L
Monochloroacetic Acid
Monobromoacetic Acid
Dichloroacetic Acid
Dalapon
Trichloroacetic Acid
Bromochloroacetic Acid
Dibromoacetic Acid
Bromodichloroacetic Acid
Chlorodibromoacetic Acid
Tribromoacetic Acid
6.00
4.00
6.00
4.00
2.00
4.00
2.00
4.00
10.0
20.0
5.24
4.36
6.89
.. 3.87
1.74
4.33
1.87
3.93
11.4
24.0
Std.
Dev.
M9/L
0.664
0.475
0.782
0.147
0.144
0.402
0.113
0.377
0.866
1.82
Rel.
Std.
Dev., %
13
11
11
3.8
8.3
9.3
6.0.
9.6
7.6
7.6
Mean
Recovery
%
87.3
109
115
96.8
87.0
108 .
93.5
98.2
114
120
Produced by the analysis of seven aliquots of fortified reagent water.
                                     552.2-29

-------
                   TABLE 5.  ANALYTE ACCURACY AND PRECISION DATAa'b
                           LEVEL 3 IN DECHLORINATED TAP  WATER0
Background
Cone.
Analyte /tg/L
Monochloroacetic Acid
Monobromoacetic Acid
Dichloroacetic Acid
Dalapon
Trichloroacetic Acid
Bromochloroacetic Acid
Dibromoacetic Acid
Bromodichloroacetic Acid
Chlorodibromoacetic Acid
Tribromoacetic Acid .
<0.6
0.420
0.625
<0.4
0.300
1.23
1.27
0.588
1.23
<2.0
Forti-
fied
Cone.
M9/L
3.00
2.00
3.00
2.00 .
1.00
2.00
1.00
2.00
5.00
10.0
Mean
Meas.
Cone.
/ig/L
2.53
2.20
3.77
1.96
1.12
2.91
2.35
2.52
6.36
11.8
Std.
Dev.
M9/L
0.090
0.034
0.096
0.157
0.167
0.062
0.110
0.388
0.502
1.65
Rel.
Std.
Dev.
%
3.6
1.5
2.5
8.0
15
2.1
4.7
15
7.9
14
Mean
Rec.
%
84.3
89.0
105
98.0
82.0
84.0
108
96.6
103
118
0 Produced  by  analysis  of seven  aliquots  of  fortified dechlorinated tap water.

b Background level  subtracted.

"Chlorinated  surface water from a  local  utility  to which  ammonium chloride was added
  as the dechlorinating agent.
                                       552.2-30

-------
                   TABLE 6.  ANALYTE ACCURACY AND PRECISION DATAa'b
                           LEVEL 3 IN HIGH IONIC STRENGTH WATERC
Background
Cone.
Analyte M9/L
Monochloroacetic Acid
Monobromoacetic Acid
Dichloroacetic Acid
Dalapon
Trichloroacetic Acid
Bromochloroacetic Acid
Dibromoacetic Acid
Bromodichloroacetic Acid
Chlorodibromoacet.ic Acid
Tribromoacetic Acid
0.761
1.47
1.50
0.675
1.01
2.06
4.36
1.07
2.48
4.63
Forti-
fied
Cone.
M9/L.
3.00
2.00
3.00
2.00
1.00
" 2.00
1.00
2.00
5.00
10.0
Mean
Meas.
Cone.
M9/L
3.32
3.19
• 4.44
2.39
1.75
3.71
5.48
3.37
7.94
17.2
Std.
Dev.
M9/L
0.429
0.099
0.264
0.259
0.110
0.269
0.255
0.308
1.00
1.55
ft '
Rel
Std
Dev
%
13
' 3.1
5.9
11
6.3
7.3
4.7
9.1
13
9.0
Mean
Rec
%
85.3
86.0
98.0
85.8
74.0
82,. 5
112
115
109,
126
a Produced  by  analysis  of seven  aliquots  of fortified  high  ionic  strength  water.

b Background level  subtracted.

c Chlorinated ground water from a water source displaying a hardness of 460 mg/L as
  CaCO,
                                        552.2-31

-------
                    TABLE 7.  ANALYTE ACCURACY AND PRECISION DATA8
                           LEVEL 3 IN HIGH HUMIC CONTENT GROUND WATERb
Background
Cone.
Analyte /ig/L
Monochloroacetic Acid
Monobromoacetic Acid
Dichloroacetic Acid
Dalapon
Trichloroacetic Acid
Bromochloroacetic Acid
Dibromoacetic Acid
Bromodichloro acetic Acid
Chlorodibromoacetic Acid
Tribromoacetic Acid
<0.6
<0.4
<0.6
<0.4
<0.2
<0.4
<0.2
<0.4
<1.0
<2.0
Forti-
fied
Cone.
3.00
2.00
3.00
2.00
1.00
2.00
1.00
2.00
5.00
10.0
Mean
Meas.
Cone.
2.91
1.99
2.88
2.00
0.618
1.82
0.715
1.99
5.50
9.67
Std.
Dev.
0.082
0.105
0.104
0.227
0.053
0.059
0.020
0.164
0.218
1.13
Rel.
Std.
Dev.
2.8
5.3
3.6
11
8.6
3.2
2.8
8.2
4.0
12
Mean
Rec.
97.0
99.5
96.0
100
61.8
91.0
71.5
99.5
110
96.7
8 Produced  by  analysis  of seven aliquots  of fortified  simulated  high  humic  content
ground water.

b Reagent water fortified at 1.0 mg/L with fulvic acid extracted from Ohio  River
  water.  Sample simulates high TOC matrix.
                                       552.2-32

-------
                    TABLE 8.   LABORATORY PERFORMANCE CHECK SOLUTION
PARAMETER
INSTRUMENT
SENSITIVITY
CHROMATOGRAPHIC
PERFORMANCE
COLUMN
PERFORMANCE
ANALYTE
MCAA
BCAA
CDBAA
SURROGATE. (2S3-DBPA)
CONC,. , pg/mL
IN MTBE
0.006
0.004
0.010
0.010
ACCEPTANCE
CRITERIA
DETECTION OF
ANALYTE;
S/Na > 3
PGFb BETWEEN
0.80 AND 1.15
RESOLUTION0
> 0.50
   S/N,  a  ratio of peak signal to baseline noise.

   peak  signal  -  measured as height of peak.
   baseline noise -  measured as maximum deviation in baseline  (in units  of  height)
                     over a width equal to the width of the base of the peak.
   PGF = Peak Gaussian  Factor

           1.83 x  W,
   PGF =
                  M/2
   where W1/2  =  the peak width at half height (in sees).
         W1/10 = the  peak width  at one-tenth height (in sees).

   This is a measure of the  symmetry of the peak.
c  Resolution between two peaks  is  defined  by the equation:

         t   '
   R =
   where t = the difference  in  elution times between the two peaks.
        Wave =  the  average peak  width of the two peaks (measurements taken  at
               baseline).

This a measure of  the  degree of separation of two peaks under specific chromatographic
conditions.
*U.S. GOVERNMENT PRINTING OFFICE: 1 995-650-006/2207 1
                                        552.2-33

-------
THIS PAGE LEFT BLANK INTENTIONALLY
             552.2-34

-------

-------

-------

-------

-------