United States
        Environmental Protection
        Agency
Office of Research and
Development
Washington, D.C. 20460
EPA/600/R-97/151
March 1998
SERA Environmental Technology
        Verification Report

        Field Portable X-ray
        Fluorescence Analyzer

        Metorex X-MET 920-MP
SUPERFUND INNOVATIVE
TECHNOLOGY EVALUATION
                                   ET
            Environmental Technology
             Verification Program
                                             070CMB98

-------
                                   EPA/600/R-97/151
                                     March 1998
Environmental Technology
Verification Report

Field Portable X-ray
Fluorescence Analyzer
Metorex X-MET 920-MP
             U.S. Environmental Protection Agency
             Office of Research and Development

-------
                                           Notice
The information in this document has been funded wholly or in part by the U.S. Environmental Protection
Agency (EPA) under Contract No. 68-CO-0047 to PRC Environmental Management, Inc. This work
supports the Superfund Innovative Technology Evaluation Program administered by the National Risk
Management Research Laboratory, Cincinnati, Ohio.  This demonstration was conducted under the
Monitoring and Measurement Technologies Program which is managed by the National Exposure Research
Laboratory-Environmental Sciences Division, Las Vegas, Nevada. It has been subjected to the Agency's
peer and administrative review, and has been approved for publication as an EPA document. Mention of
corporation names, trade names, or commercial products does not constitute endorsement or
recommendation for use of specific products.

-------
                     UNITED STATES ENVIRONMENTAL PROTECTION AGENCY
                                    Office of Research and Development
                                        Washington, D.C. 20460

             ^
   % ^M|u^ -^                                                               ENVIRONMENTAL TECHNOLOGY
    <*..      _


-------
discussion of results, may be found in the report entitled "Environmental Technology Verification Report, Field
Portable X-ray Fluorescence Analyzer, Metorex X-MET 920-MP." The EPA document number for this report is
EPA/600/R-97/151.

The EPA SW-846 Method 6200 was tested and validated using the data derived from this demonstration.  This
method may be used to support the general application of FPXRF for environmental analysis.

TECHNOLOGY DESCRIPTION

This analyzer  operates on  the principle of energy dispersive X-ray  fluorescence spectroscopy where the
characteristic energy components of the excited X-ray spectrum are analyzed directly by an energy proportional
response in an X-ray detector.  Energy dispersion affords a highly efficient, full-spectrum measurement which
enables the use of low intensity excitation sources (such as radioisotopes) and compact battery-powered, field-
portable electronics.  The FPXRF instruments are designed to provide  rapid analysis of metals in soil.  This
information allows investigation and remediation decisions to be made on-site and reduces the number of samples
that need to be submitted for laboratory analysis. In the operation of these instruments, the user must be aware that
FPXRF analyzers do not respond well to chromium and that field detection limits may be 5 to 10 times greater than
conventional laboratory methods.  As with all field collection programs, a portion of the samples should  be sent
to a laboratory for confirmatory analyses.

The X-MET 920-MP is designed to produce quantitative data on metals in soils, sludges, and other solids.  The X-
MET 920-MP consists of a laptop computer, an electronics unit, and a surface analysis probe system (SAPS). The
electronics is housed in a rugged, weatherproof, self-contained case, weighing about 5 pounds that can be operated
from battery power up to 8 hours. The SAPS is designed to house one excitation source (cadmium-109  for this
demonstration) and a gas-filled proportional counter detector. The SAPS weighs about 3 pounds and is specifically
designed for in situ analysis, but can be adapted for measurement of samples in cups. The single excitation source
limits the number of metals that can be quantified.  The X-MET 920-MP is operated and calibrated using the "X-
MET"  software to analyze samples with an empirical calibration.  Training and field experience is necessary to
successfully derive empirical calibration curves and to operate the "X-MET" software.  The X-MET 920-MP
reported the analytes arsenic, barium, copper, chromium, lead, nickel, and zinc for this demonstration using source
count times between 30 and 180 seconds. At the time of the demonstration, the cost of the X-MET 920-MP with
the SAPS probe and cadmium-109 source (including the laptop computer) was $36,325, or it could be leased for
$3,633 per month.

VERIFICATION OF PERFORMANCE

The performance characteristics of the X-MET 920-MP include the following:

•  Detection limits: Precision-based detection limits were determined by collecting  10 replicate measurements
   on site-specific soil samples with metals concentrations 2 to 5 times the expected MDLs. The results were less
   than or equal to 50 milligrams per kilogram (mg/kg) for all analytes except barium (330 mg/kg) and chromium
   (115 mg/kg). Barium is normally analyzed using an americium-241 source; therefore, its detection limit was
   expected to be high. A value for nickel could not be determined because the soil concentration of this analyte
   was too low.
•  Throughput:  Average throughput was found to be between 8 and 14 analyses per hour, depending on count
   times. This rate only represents the analysis time since different personnel were used to prepare the samples.
•  Drift:  Based on an evaluation of results  from periodic analysis of a site-specific control sample, with a few
   exceptions, drift was -15 to +15 percent.  Lead and arsenic displayed the least drift at both sites.
•  Completeness: The X-MET 920-MP produced results for  1,168 of the  1,260 samples for a completeness of
   92.7 percent. This was less than the demonstration objective of 95 percent.  Operator error  and computer
   software and hardware problems reduced completeness. None of the data loss was caused by mechanical or
   electronic malfunctions of the analyzer.
EPA-VS-SCM-08                  The accompanying notice is an integral part of this verification statement                  March 1998

                                                 iv

-------
•  Blank results: The X-MET 920-MP reported values for arsenic and copper above the precision-based method
   detection limits at the ASARCO site and values for chromium, lead, and zinc above the MDL at the RV Hopkins
   site.  Analyzer blanks were composed of a pure lithium carbonate that was processed using the  sample
   preparation steps.
•  Precision:  The goal of the demonstration was to achieve relative standard deviations (RSD) less than 20
   percent at analyte concentrations of 5 to 10 times the method detection limit.  The RSD values for the reported
   analytes were less than 8 percent.  Chromium and nickel were not determined due to a lack of sufficient data
   in the specified concentration range.
•  Accuracy:  Intramethod accuracy was assessed  using site-specific soil PE samples. The results showed that
   7 of 32 (21.9 percent) of the PE sample analytes had recoveries within a quantitative acceptance range of the
   80 - 120 percent.
•  Comparability: This demonstration showed thatthe X-MET 920-MP produced data that exhibited a Iog10-log10
   linear correlation to the reference data. The coefficient of determination (r2) which is a measure of the degree
   of correlation between the reference and field data was 0.95 for arsenic, 0.88 for lead, 0.69 for copper, 0.68 for
   chromium, and 0.55 for zinc.  Using data from the RV Hopkins clay soil produced values of 0.62 for barium and
   0.32 for nickel.
•  Data quality  levels:   Using  the demonstration derived precision RSD results and the  coefficient  of
   determination as the primary qualifiers, the X-MET 920-MP produced definitive level data for arsenic and lead
   and data of qualitative screening level for copper, barium, and zinc. No recommendation regarding data quality
   for chromium or nickel could be made due to a lack of precision or comparability data.

The results of the demonstration show that the Metorex  X-MET 920-MP can provide useful, cost-effective data
for environmental problem-solving and decision-making.  Undoubtedly, it will be employed  in a variety  of
applications, ranging from serving as a complement to data generated in a fixed analytical laboratory to generating
data that will stand alone in the decision-making  process.   As with any technology selection, the user must
determine what is appropriate for the application and the project  data quality objectives.
Gary J. Foley, Ph.D.
Director
National Exposure Research Laboratory
Office of Research and Development
  NOTICE:  EPA verifications are based on an evaluation of technology performance under specific, predetermined criteria and the
  appropriate quality assurance procedures. EPA makes no expressed or implied warranties as to the performance of the technology
  and does not certify that a technology will always, under circumstances other than those tested, operate at the levels verified. The
  end user is solely responsible for complying with any and all applicable Federal, State, and Local requirements.
EPA-VS-SCM-08                    The accompanying notice is an integral part of this verification statement                   March 1998

                                                    V

-------
                                          Foreword
The U.S. Environmental Protection Agency (EPA) is charged by Congress with protecting the Nation's
land, air, and water resources. Under a mandate of national environmental laws, the Agency strives to
formulate and implement actions leading to a compatible balance between human activities and the ability
of natural systems to support and nurture life. To meet this mandate, the EPA's Office of Research and
Development (ORD) provides data and science support that can be used to solve environmental problems
and to build the scientific knowledge base needed to manage our ecological resources wisely, to understand
how pollutants affect our health, and to prevent or reduce environmental risks.

The National Exposure Research Laboratory (NERL) is the Agency's center for the investigation of
technical and management approaches for identifying and quantifying risks to human health and the
environment.  Goals of the Laboratory's research program are to develop and evaluate technologies for the
characterization and monitoring of air, soil, and water; support regulatory and policy decisions; and
provide the science support needed to ensure effective implementation of environmental regulations and
strategies.

The EPA's Superfund Innovative Technology Evaluation (SITE) Program evaluates technologies for the
characterization and remediation of contaminated Superfund and Resource Conservation and Recovery Act
(RCRA)  corrective action sites.  The SITE Program was created to provide reliable cost and performance
data to speed the acceptance of innovative characterization  and monitoring technologies.

Effective measurement and monitoring technologies are needed to assess the degree of contamination at a
site, to provide data which can be used to determine the risk to public health or the environment, to supply
the necessary cost and performance data to select the most appropriate technology, and to monitor the
success or failure of a  remediation process.  One component of the SITE Program, the Monitoring and
Measurement Technologies  Program, demonstrates and evaluates innovative technologies to meet these
needs.

Candidate technologies can originate from within the federal government or from the private sector.
Through  the SITE Program, developers are given the opportunity to conduct a rigorous demonstration of
their technology's performance under realistic field conditions. By completing the evaluation and
distributing the  results, the Agency establishes a baseline for acceptance and use of these technologies. The
Monitoring and Measurement Technologies Program is managed by ORD's  Environmental Sciences
Division  in Las  Vegas, Nevada.
                                     Gary J. Foley, Ph.D.
                                     Director
                                     National Exposure Research Laboratory
                                     Office of Research and Development
                                               VI

-------
                                          Abstract
In April 1995, the U.S. Environmental Protection Agency (EPA) conducted a demonstration of field
portable X-ray fluorescence (FPXRF) analyzers. The primary objectives of this demonstration were (1) to
determine how well FPXRF analyzers perform in comparison to standard reference methods, (2) to identify
the effects of sample matrix variations on the performance of FPXRF, (3) to determine the logistical and
economic resources needed to operate these analyzers, and (4) to test and validate an SW-846 draft method
for FPXRF analysis.  The demonstration design was subjected to extensive review and comment by the
EPA's National Exposure Research Laboratory, EPA Regional and Headquarters Superfund technical
staff, the EPA's Office of Solid Waste-Methods Section, and the technology developers.

Two sites were used for this demonstration:  the RV Hopkins site and the ASARCO Tacoma Smelter site
(ASARCO). RV Hopkins is an active steel drum recycling facility and a former battery recycling
operation. It is located in Davenport, Iowa.  The ASARCO site is a former copper and lead smelter and is
located in Tacoma, Washington.  The test samples analyzed during this demonstration were evenly
distributed between three distinct soil textures:  sand, loam, and clay.  The reference methods used to
evaluate the comparability of data were EPA SW-846 Methods 3050A and 6010A, "Acid Digestion of
Sediments,  Sludges, and Soils" and "Inductively Coupled Plasma-Atomic Emission Spectroscopy,"
respectively.

The FPXRF analyzers tested in this demonstration were designed to provide rapid, real-time analysis of
metals concentrations in soil samples.  This information will allow investigation and remediation decisions
to be made  on-site more efficiently and can reduce the number of samples that need to be submitted for
confirmatory analysis. Of the seven commercially available analyzers tested, one is manufactured by Niton
Corporation (the XL Spectrum Analyzer); two are manufactured by TN Spectrace (the TN 9000 and TN
Pb Analyzer); two are manufactured by Metorex Inc. (the X-MET 920-P Analyzer and the X-MET 920-
MP Analyzer); one is manufactured by HNU Systems, Inc. (the SEFA-P Analyzer); and one is
manufactured by Scitec Corporation (the MAP Spectrum Analyzer).  The X-MET 940, a prototype
FPXRF analyzer developed by Metorex, was given special consideration and replaced the X-MET 920-P
for a portion of the demonstration.  This environmental technology verification report (ETVR) presents
information relative to the Metorex X-MET 920-MP. Separate ETVRs have been published for the other
analyzers demonstrated.

Approximately three days of operational downtime was experienced by the analyzer due to computer
software and hardware problems.  Most of these problems were due to operator error or inexperience.
None of the downtime or data loss was associated with mechanical or electronic malfunctions of the
analyzer. Quantitative data was provided by the analyzer on a real-time basis. The X-MET 920-MP
Analyzer reported arsenic, chromium, copper, lead, zinc, nickel, and barium.  This analyzer used count
times ranging from 30 live-seconds for in s/YH-unprepared samples at the ASARCO site to 180 live-seconds
for intrusive-prepared samples at the RV Hopkins  site. These count times resulted in a  sample throughput
averaging between 8 and 14 samples per hour.  The X-MET 920-MP Analyzer provided definitive data

                                              vii

-------
(equivalent to reference data) for arsenic and lead; and qualitative screening level data (identifies the
presence or absence of a contaminant) for copper, barium, and zinc.  Insufficient precision data precluded
an assignment of data quality levels for nickel or chromium.

This study showed that the analyzer produced data that exhibited a Iog10-log10 linear correlation to the
reference data.  The analyzer generally exhibited a precision similar to the reference methods. The analyzer
exhibited precision of less than 10 percent relative standard deviation at 5 to 10 times the method detection
limit (MDL) for all of the reported analytes except chromium and nickel. The precision evaluation was
confounded by changing count times.  The precision study indicated that count times probably had no effect
on the precision for all target analytes except copper and lead. For copper and lead, the increasing count
times caused a 2- to a 10-fold increase in precision.  The analyzer's quantitative results were based on an
empirical calibration using site-specific calibration samples.

This demonstration found that the X-MET 920-MP Analyzer was generally simple to operate in the field;
however, its physical configuration made it more practical for use as a benchtop unit.  The auxiliary
computer and cumbersome power requirements of commercial laptop computers limited its utility as an in
situ instrument.  The operator required no specialized experience or training for normal operation of the
analyzer. However, ownership and operation of this analyzer may require specific licensing by a state
nuclear regulatory agency. There are specific radiation safety training requirements and costs associated
with this type of license.

The Metorex X-MET 920-MP Analyzer can provide rapid, real-time analysis of the metals content of soil
samples at hazardous waste sites. The analyzer can quickly identify contaminated areas from
noncontaminated areas allowing investigation and remediation decisions to be made more efficiently on-
site and reduce the number of samples that need to be submitted for confirmatory analysis.
                                                VIII

-------
                               Table of Contents

Section                                                                       Page

Notice	  ii
Verification Statement	  iii
Foreword	  vi
Abstract  	vii
List of Figures	  xi
List of Tables 	xii
List of Abbreviations and Acronyms	  xiii
Acknowledgments	xv

1  Executive Summary	  1

2  Introduction	  3
    Demonstration Background, Purpose, and Objectives  	  3
    Reference Methods	  4
    Site Selection	  5
    Predemonstration Sampling  	  7
    Experimental Design	  8
    Qualitative Factors	  10
    Quantitative Factors 	  10
    Evaluation of Analyzer Performance	12
    Deviations from the Demonstration Plan	19
    Sample  Homogenization	21

3  Reference Laboratory Results  	22
    Reference Laboratory Methods	22
    Reference Laboratory Quality Control	23
    Quality Control Review of Reference Laboratory Data  	24
       Reference Laboratory Sample Receipt, Handling, and Storage Procedures  	24
       Sample Holding Times  	25
       Initial and Continuing Calibrations	25
       Detection Limits	25
       Method Blank Samples	26
       Laboratory Control Samples	26
       Predigestion Matrix Spike Samples	26
       Postdigestion Matrix Spike Samples	27
       Predigestion Laboratory Duplicate Samples	28
       Postdigestion Laboratory Duplicate Samples  	28
       Performance Evaluation Samples	28
       Standard Reference Material Samples	29
       Data  Review, Validation, and Reporting 	29

                                         ix

-------
Section                                                                      Page

    Quality Assessment of Reference Laboratory Data	30
       Precision	30
       Accuracy	31
       Representativeness  	33
       Completeness	33
       Comparability  	36
    Use of Qualified Data for Statistical Analysis	37

4 X-MET 920-MP Analyzer 	40
    Theory of FPXRF Analysis  	40
    Background	41
    Operational Characteristics	42
       Equipment and Accessories  	42
       Operation of the Analyzer	44
       Background of the Technology Operators	45
       Training	46
       Reliability	46
       Health and Safety	47
       Cost	48
    Performance Factors	49
       Detection Limits	49
       Throughput	50
       Drift	51
    Intramethod Assessment	52
       Blanks	52
       Completeness	52
       Precision	52
       Accuracy	54
    Intermethod Assessment	57

5 Applications Assessment and Considerations  	66
    General  Operational Guidance  	70

6 References	72

-------
                                 List of Figures
2-1  Sample Preparation and Analysis	9
2-2  Linear and Log-log Data Plots	14
3-1  Pre- and Postdigestion Duplicate Samples	31
3-2  Reference Method PE and CRM Results  	34
3-3  Reference Method SRM Results	38
4-1  Principle of Source Excited X-ray Fluorescence  	41
4-2  Critical Zone for the Determination of a Field-based Method Detection Limit for Copper  50
4-3  Drift Summary	51
4-4  Site-specific  PE Sample Results	56
4-5  Sample Preparation Effects on Arsenic Results  	60
                                         XI

-------
                                  List of Tables
Table
2-1    Performance and Comparability Variables Evaluated	11
2-2    Criteria for Characterizing Data Quality  	16
3-1    Reference Laboratory Quality Control Parameters	23
3-2    SW-846 Method 6010A LRLs for Target Analytes	26
3-3    Reference Laboratory Accuracy Data for Target Analytes	32
3-4    SRM Performance Data for Target Analytes 	36
3-5    Leach Percent Recoveries for Select NIST SRMs 	37
4-1    Analyzer Instrument Specifications	43
4-2    Instrument and  Field Operation Costs  	49
4-3    Method Detection Limits	50
4-4    Precision Summary	53
4-5    Precision as Affected by Sample Preparation and Count Times  	54
4-6    Accuracy Summary for Site-Specific PE Results  	55
4-7    Regression Parameters by Primary Variable 	58
4-8    Regression Parameters for the Preparation Variable Sorted by Soil Texture	61
4-9    Regression Parameters for the Preparation Variable Sorted by Site Name	63
4-10  Summary of Data Quality Level Parameters	65
5-1    Summary of Test Results and Operational Features 	67
5-2    Effects of Data  Correction on FPXRF Comparability to Reference Data for All
         In Situ-Prepared Samples	68
                                         XII

-------
                     List of Abbreviations and Acronyms
a
P
Am241
CCB
CCV
Cd109
Cl
CLP
cm
cm2
cm3
Co57
CRM
DC
EPA
ERA
ETVR
eV
FPXRF
ICAL
ICB
ICP-AES
ICS
ICV
IDL
IDW
keV
LCD
LCS
lOQio
LRL
MCA
mCi
MDL
mg/kg
mL
mm
MMTP
mrem/hr
MRI
NERL-ESD
alpha
beta
americium-241
continuing calibration blank
continuing calibration verification
cadmium-109
confidence interval
Contract Laboratory Program
centimeter
centimeter squared
cubic centimeter
cobalt-57
certified reference material
direct current
Environmental Protection Agency
Environmental Resource Associates
environmental technology verification report
electron volt
field portable X-ray fluorescence
initial calibration
initial calibration blank
inductively coupled plasma-atomic emission spectroscopy
interference check standard
initial calibration verification
instrument detection limit
investigation-derived waste
kiloelectron volt
liquid crystal  display
laboratory control samples
base 10 logarithm
lower reporting limit
multichannel analyzer
millicurie
method detection limit
milligram  per kilogram
milliliter
millimeter
Monitoring and Measurement Technologies Program
millirems  per hour
Midwest Research Institute
National Exposure Research Laboratory—Environmental Sciences Division
                                        XIII

-------
NIST
OSW
PAL
PARCC
PC
PE
PI
ppm
PRC
QA
QAPP
QC
r
r2
RCRA
RPD
RSD
RTC
SD
SITE
SOP
SRM
SSCS
TC
USGS
XRF
National Institute of Standards and Technology
Office of Solid Waste
performance acceptance limit
precision, accuracy, representativeness, completeness, and comparability
personal computer
performance evaluation
prediction interval
part per million
PRC Environmental Management, Inc.
quality assurance
quality assurance project plan
quality control
correlation coefficient
coefficient of determination
Resource Conservation and Recovery Act
relative percent difference
relative standard deviation
Resource Technology Corporation
standard deviation
Superfund Innovative Technology Evaluation
standard operating procedure
standard reference material
site-specific calibration sample
toxicity characteristic
United States Geological Survey
X-ray fluorescence
                                        XIV

-------
                                  Acknowledgments
The U.S. Environmental Protection Agency (EPA) wishes to acknowledge the support of all those who
helped plan and conduct this demonstration, interpret data, and prepare this report.  In particular, for
demonstration site access and relevant background information, Tom Aldridge (ASARCO) and Harold
Abdo (RV Hopkins); for turnkey implementation of this demonstration, Eric Hess, Patrick Splichal, and
Harry Ellis (PRC Environmental Management, Inc.); for editorial and publication support, Suzanne
Ladish, Anne Witebsky, Karen Bellinger, and Ed Hubert (PRC  Environmental Management, Inc.); for
technical and peer review, Paula Hirtz, David Farnam, and Alan Byrnes (PRC Environmental
Management, Inc.); for analyzer operation, Wes McCall (PRC Environmental Management, Inc.); for
sample preparation, Scott Schulte, Keith Brown, and Curt Enos (PRC Environmental Management, Inc.);
for EPA project management, Stephen Billets, National Exposure Research Laboratory-Environmental
Sciences Division; and for peer review, Sam Goforth (independent consultant), John Wallace (Wallace
Technologies),  and Shirley Wasson (National Risk Management Research Laboratory). In addition, we
gratefully acknowledge the participation of Oliver Fordham, EPA Office of Solid Waste; Piper Peterson,
EPA Region 10; Brian Mitchell, EPA Region 7; and James R. Pasmore, Metorex, Inc.
                                             xv

-------
                                         Section 1
                                  Executive Summary
    In April 1995, the U.S. Environmental Protection Agency (EPA) sponsored a demonstration of field
portable X-ray fluorescence (FPXRF) analyzers.  The primary objectives of this demonstration were to
evaluate FPXRF analyzers for: (1) their analytical performance relative to standard analytical methods,
(2) the influence of sample matrix variations (texture, moisture, heterogeneity, and chemical composition)
on performance, (3) the logistical and economical resources needed to operate these technologies in the
field, and (4) to test and validate an SW-846 draft method for FPXRF analysis. Secondary objectives for
this demonstration were to evaluate FPXRF analyzers for their reliability, ruggedness, cost, range of
usefulness, data quality, and ease of operation.

    This demonstration was intended to provide users with a reference measure of performance and to act
as a guide for the application of this technology. In this demonstration, the reference methods for
evaluating the comparability of data were SW-846 Methods 3050A and 6010A, "Acid Digestion of
Sediments, Sludges, and Soils" and "Inductively Coupled Plasma-Atomic Emission Spectroscopy (ICP-
AES)," respectively.

    The EPA requested that PRC Environmental Management, Inc. (PRC) assist in the planning,
execution, and reporting of a demonstration of FPXRF analyzers. This demonstration was conducted
under the EPA's Superfund Innovative Technology Evaluation (SITE) Program and managed by the
National Exposure Research Laboratory-Environmental Sciences Division (NERL-ESD) under the
Monitoring and Measurement Technologies Program (MMTP), Las Vegas, Nevada.

    The FPXRF analyzers tested in this demonstration were designed to provide rapid, real-time analysis of
metals concentrations in soil samples. This information will allow investigation and remediation decisions
to be made on-site more efficiently, and it should reduce the number of samples that need to be submitted
for confirmatory analysis.  Of the seven commercially available analyzers evaluated, one is manufactured
by Niton Corporation (the Niton XL Spectrum Analyzer); two are manufactured by Metorex Inc.  (the X-
MET 920-P Analyzer and the X-MET 920-MP Analyzer); two are manufactured by TN Spectrace (the TN
9000 and the TN Pb Analyzer); one is manufactured by HNU Systems, Inc. (the SEFA-P Analyzer); and
one is manufactured by Scitec Corporation (the MAP Spectrum Analyzer). The X-MET 940, a prototype
FPXRF analyzer developed by Metorex, was given special consideration and replaced the X-MET 920-P
for part of the RV Hopkins sample analyses. This environmental technology verification report (ETVR)
presents information relative to the Metorex X-MET 920-MP Analyzer. Separate ETVRs will be
published for the other analyzers that were demonstrated.

    The target analytes for this demonstration were selected from the Resource Conservation and Recovery
Act's (RCRA) Toxicity Characteristic (TC) list, analytes known to have a high aquatic toxicity and

-------
analytes likely to produce interferences for the FPXRF analyzers. The primary analytes for these
comparisons were arsenic, barium, chromium, copper, lead, and zinc; nickel, iron, cadmium, and antimony
were secondary analytes.

    To demonstrate these analyzers, hazardous waste sites in Iowa (the RV Hopkins site) and in the State
of Washington (the ASARCO site) were selected.  These sites were selected because they exhibit a wide
range of concentrations for most of the target analytes, are located in different climatological regions of the
United States, and combined they exhibit three distinct soil textures: sand, loam, and clay.

    This demonstration found that the X-MET 920-MP Analyzer was generally simple to operate in the
field.  However, unexpected computer software and hardware problems caused data loss and project
downtime.  In addition, the physical configuration of the analyzer made its use in the in situ mode
cumbersome. Its physical configuration is more adapted to a benchtop application.  The developer
provided a training course for the operator that was similar to that provided to a purchaser of the
equipment.  The training encompassed enough FPXRF theory and hands-on analyzer use to allow the
operator to  calibrate the analyzer, manipulate  the data collection software, and adjust instrument
parameters  such as count times and target analytes. Metorex provided accessible and timely field support.
The analyzer was portable and capable of operating continuously over a 12-hour work day with
appropriate battery changes.  The data collection and interpretation software is designed to operate with an
auxiliary computer system. A laptop computer was used. The field portability was reduced by the size and
power requirements of the laptop computer. The computer could only operate 1.5 to 2 hours on a battery.
The environmental conditions encountered at the ASARCO site caused no operational downtime for the
analyzer.

    The analyzer used one radioactive source  coupled with a gas proportional detector. The type and
strength of the analyzer's source allow it to produce reliable data at count times as short as 30 live-seconds.
The count times used in this demonstration resulted in a sample throughput averaging 8-14 samples per
hour.

    The X-MET 920-MP Analyzer produced data meeting definitive level criteria (equivalent to reference
data) for arsenic and lead; and produced qualitative screening level data (identifies the presence or absence
of a contaminant) for barium, zinc, and copper. Data quality levels for nickel and chromium were not
assigned due to lack of sufficient data.

    The analyzer exhibited a similar precision relative to the reference methods.  The chromium and nickel
data generally showed the poorest precision of the reported analytes. No values were reported for these
analytes at the  5 to 10 times MDL range. Field-based method detection limits (MDL) for this analyzer
were higher than precision-based or developer-provided MDLs. However, this relationship was
confounded by increased count times with increasing sample preparations. Site and soil texture did not
appear to affect data comparability. The probe's high counting efficiency and the empirical calibration
seemed to minimize any impact of inter-element interferences.

    Based on the performance of the analyzer, this demonstration found it to be a useful tool for
characterizing the concentration of select metals in environmental soil samples. As with all FPXRF
analyzers, unless a user has regulatory approval, confirmatory (reference) sampling and data correction is
recommended when using this technology for  site characterization or remediation monitoring.

-------
                                         Section 2
                                        Introduction
    This environmental technology verification report (ETVR) presents information from the demonstration
of the X-MET 920-MP Analyzer. This analyzer was developed by Metorex, Inc. to perform elemental
analyses (metals quantitation) in the field. This analyzer uses a gas-filled proportional detector and
cadmium-109 (Cd109) source to quantitate metals concentrations. The analyzer can be operated in either an
in situ or intrusive mode. The in situ mode, commonly called a "point-and-shoot" mode, requires that the
point of measurement on the soil surface is cleared of loose debris and organic matter, the analyzer's probe
is placed in direct contact with the soil surface, and a measurement is taken. In the intrusive mode  of
operation, a soil sample is physically collected, dried or sieved, and then placed in a sample cup. The
sample cup is placed into an analysis chamber on the probe and a measurement is taken.

    This section summarizes general information about the demonstration such as the purpose, objectives,
and design. Section 3 presents and discusses the quality of the data produced by the reference methods
against which the analyzer was evaluated. Section 4 discusses the  X-MET 920-MP Analyzer, capabilities,
equipment, accessories, accuracy, precision, comparability to reference methods, and other evaluation
factors.  Section 5 discusses potential applications of the analyzer, presents a method for data correction,
and suggests a framework for a standard operating procedure (SOP). Section 6 lists the references cited in
this ETVR.

Demonstration Background,  Purpose,  and Objectives

    The demonstration was conducted under the Monitoring and Measurement Technologies Program
(MMTP), a component of the SITE Program. MMTP is managed by NERL-ESD, Las Vegas, Nevada.
The goal of the MMTP is to identify and demonstrate new, innovative, and commercially available
technologies that can sample, identify, quantify, or monitor changes in contaminants at hazardous waste
sites.  This includes those technologies that can be used to determine the physical characteristics of a site
more economically, efficiently, and safely than conventional technologies.  The SITE Program is
administered by the National Risk Management Research Laboratory, Cincinnati, Ohio.

    The purpose of this demonstration was to provide the information needed to fairly and thoroughly
evaluate how well FPXRF analyzers identify and quantify concentrations of metals in soils. The primary
objectives were to evaluate FPXRF analyzers in the  following areas: (1) their accuracy and precision
relative to conventional analytical methods; (2) the influence of sample matrix variations (texture, moisture,
heterogeneity, and chemical composition) on performances; (3) the  logistical and economic resources
needed to operate these analyzers; and (4) to test and validate an SW-846 draft method for FPXRF
analysis.

-------
    Secondary objectives for this demonstration were to evaluate FPXRF analyzers for their reliability,
ruggedness, cost, range of usefulness, data quality, and ease of operation.  The performances of the FPXRF
analyzers were not compared against each other. Instead, the performance of each analyzer was
independently and individually compared to that of standard conventional analytical methods commonly
used in regulatory enforcement or compliance activities. In addition, the analyzer's performances were
assessed relative to measurements of standard reference materials (SRM), performance evaluation (PE)
samples, and other quality control (QC) samples.

    A special request was made by Mr. Oliver Fordham, the project's technical advisor, EPA Office of
Solid Waste (OSW), for Midwest Research Institute (MRI) to analyze some of the soil samples to validate
the performance of draft Method 3052 "Microwave Assisted Acid Digestion of Ash and Other Siliceous
Wastes." Thirty percent of the soil samples were extracted using draft Method 3052 and then analyzed by
Method 6010A. The data generated from the draft Method 3052 and Method 6010A analysis were not
used for comparative purposes to the FPXRF data in this demonstration.

Reference Methods

    To assess the performance of each analyzer, FPXRF data was compared to reference data. The
reference methods used for this assessment were EPA SW-846 Methods 3050A/6010A, which are
considered the standards  for metals analysis in soil for environmental applications.  For purposes of this
discussion, the term "reference" was  substituted for "confirmatory" since the data was used as a baseline
for comparison. In accordance with Federal Acquisition Regulations, MRI was awarded a subcontract to
analyze soil samples using the reference methods.  The award was made based on MRI's costs, ability to
meet the demonstration's quality assurance project plan (QAPP), requirements, and as the only commercial
laboratory identified that could perform all the analyses in the required timeframe.

    Method 3050A is the standard acid extraction procedure for determining metals concentrations in soil
samples.  It is not a total digestion method, and it potentially does not extract all the metals in a soil sample.
Method 6010A is the standard method used to analyze Method 3050A extracts. Both of these methods are
described in Section  3.

    High quality, well documented reference laboratory results were essential for meeting the objectives of
the demonstration. For an accurate assessment, the reference methods have to provide a known level of
data quality. For all measurement and monitoring activities conducted by the EPA, the Agency requires
that data quality parameters be established based on the end use of the data. Data quality parameters
usually include five indicators of data quality often known as the PARCC parameters:  precision, accuracy,
representativeness, completeness, and comparability.  In addition, method detection limits (MDLs) are also
used to assess data quality.

    Reference methods were evaluated using the PARCC parameters to establish the quality of data
generated and to ensure that the comparison of FPXRF analyzers to reference methods was acceptable.
The following narrative provides definitions of each of the PARCC parameters.

    Precision refers to the degree of mutual agreement between replicate measurements and provides an
estimate of random error.  Precision is often expressed in terms of relative standard deviation (RSD)
between replicate samples. The term relative percent difference  (RPD) is used  to provide an estimate of
random error between duplicate samples.

-------
    Accuracy refers to the difference between a sample result and the reference or true value.  Bias, a
measure of the departure from perfect accuracy, can be calculated from the reference or true value.
Accuracy and bias for the reference laboratory were assessed by evaluating calibration standard linearity,
method blank results and the percent recoveries of matrix spike samples, laboratory control samples (LCS),
standard reference materials (SRMs), and PE samples.

    Representativeness refers to the degree to which data accurately and precisely measures the conditions
and characteristics of the parameter of interest. Representativeness for the reference laboratory was
ensured by executing  consistent sample collection procedures including sample locations, sampling
procedures, storage, packaging, shipping, equipment decontamination, and proper laboratory sample
handling procedures.  Representativeness was ensured by using the  appropriate reference method to provide
results that reflect the most accurate and precise measurement it was capable of achieving. The
combination of existing method requirements supplemented by the demonstration QAPP provided the
guidance to assure optimum performance of the method. Representativeness was assessed by evaluating
calibration standards, method blank samples, duplicate samples, and PE samples.

    Completeness refers to the amount of data collected from a measurement process compared to the
amount that was expected to be obtained. For the reference data, completeness referred to the  proportion of
valid, acceptable data generated.

    Comparability refers to the confidence with which one data set  can be compared to another.  Data
generated from the reference methods should provide comparable data to any other laboratory performing
analysis of the same samples with the same analytical methods. Comparability for the reference methods
was achieved through the use of standard operating  procedures (SOPs), EPA-published guidance, and the
demonstration QAPP. QC samples that were used to evaluate comparability include: calibration
standards, method blank samples, matrix spike samples, replicate samples, LCSs, SRMs, and PE samples.

Site  Selection

    PRC conducted a search for suitable demonstration sites between September and November 1994.  The
following criteria were used to select appropriate sites:

    •  The site  owner had to agree to allow access for the demonstration.

    •  The site  had to have soil contaminated with some or all of the target heavy metals. (Slag, ash, and
      other deposits of mineralized metals would not be assessed during the demonstration.)

    •  The site  had to be accessible to two-wheel drive vehicles.

    •  The site  had to exhibit one or more of the following soil textures: sand, clay, or loam.

    •  The site  had to exhibit surface soil contamination.

    •  The sites had to be situated in different climatological environments.

    PRC contacted NERL-ESD, regional EPA offices, state environmental agencies, metals fabrication,
and smelting contacts to create an initial list of potential demonstration sites. PRC received considerable
assistance from the EPA RCRA and Superfund Branches in Regions 4, 6, 7, 8, 9, and 10.  PRC also

-------
contacted the Montana Department of Health and Environment, the Nevada Bureau of Mines and Geology,
the Oklahoma Department of Environmental Quality, the Arizona Department of Environmental Quality,
the Missouri Department of Natural Resources, the Arizona Bureau of Geology, and the New Mexico
Bureau of Mines and Mineral Resources. PRC surveyed its offices in Kansas City, Kansas; Atlanta,
Georgia; Denver, Colorado; Dallas, Texas; Albuquerque, New Mexico; Helena, Montana; Chicago,
Illinois; Seattle, Washington; and San Francisco, California, for information regarding potential sites.
These PRC offices have existing RCRA, Superfund, or Navy environmental contracts that allow access to
regional, state, and federal site information. PRC also used the Record of Decision Scan database (Morgan
and others 1993) to search for appropriate sites.

    PRC screened 46 potential sites based on the site-selection criteria with the  assistance of the various
contacts listed above. Based on this screening effort, PRC and the EPA determined that the RV Hopkins
and ASARCO sites met most of the site-selection criteria, and therefore, would  be acceptable for the
demonstration.

    The ASARCO site consists of 67 acres of land adjacent to Commencement  Bay.  The site is marked by
steep slopes leading into the bay, a slag fill that was used to extend the original  shoreline, a cooling water
pond, and various buildings associated with the smelting process. Partial facility demolition was conducted
in 1987. Most of the buildings were demolished between 1993 and 1994.  The only buildings remaining
are the Fine Ore Building, the Administrative Building, and a Maintenance Garage.

    Past soil sampling results have targeted four general areas of the site as acceptable candidates for the
demonstration:  the plant administration area, the former cooling pond, the  1987 demolition area, and
certain off-site residential areas adjacent to the smelter stack.  Previous sampling has  shown surficial soils
to be more contaminated than subsurface soils. Arsenic, copper, and lead are the predominant contaminants
in the local soils. The highest arsenic concentrations were found in the soils around the former arsenic
kitchen, along with cadmium and mercury.  The soils around the former cooling pond contained the highest
copper concentrations and high levels of silver, selenium, barium, and chromium.  Lead concentrations are
highest northeast of the arsenic plant.

    Much of the smelter site is covered with artificial fill material of varying thickness and composition.
Two general types of fill are found on-site:  granular and slag.  The composition of the granular fill
material ranges from sand to silt with demolition debris and slag debris intermixed throughout.  The slag
fill is a solid, fractured media restricted to the plant site.  The surface soil in the plant administration area
has a layer of slag particles on top, ranging from  1 to 3 inches thick. Surficial material in the parking lot
area and southwest of the stack is mostly of glacial origin and composed of various mixtures of sand,
gravel, and cobbles.  The soils around the former cooling pond are fine-grained  lacustrine silts and clays.
Alluvium upgradient of the former cooling pond has been almost entirely covered with granular fill
material. Generally, soils in the arsenic kitchen and stack hill areas are sand mixed with gravel or sandy
clay mixed with cobbles.  No slag was analyzed as part of this demonstration.

    The RV Hopkins site is located in the west end of Davenport, Iowa. The facility  occupies
approximately 6.7 acres in a heavy industrial/commercial zoned area. Industrial activities in the area of the
RV Hopkins property included the manufacture of railroad locomotive engines during the mid-1800's. The
RV Hopkins property was a rock quarry during the late  1800's. Aerial surveys  beginning in 1929 show
that the rock quarry occupied the majority of the site initially, gradually decreasing until it was completely
filled by 1982.  It was reported that the site was used to dispose of demolition debris, automotive, and scrap
metal.  The site also has been used by a company that recycled lead acid batteries.

-------
    RV Hopkins began operating as a drum reconditioner in 1951 across the street from its current
location.  In 1964, the site owner reportedly covered the former quarry area of the site with foundry sand.
No foundry sand was analyzed as part of this demonstration. RV Hopkins receives between 400 and 600
drums per day for reconditioning, accepting only drums that meet the definition of "empty" according to 40
Code of Federal Regulations 261.7.  Most of the drums received at the facility come from the paint, oil, and
chemical industries. The surrounding area is reported to be underlain by Devonian-aged Wapsipinicon
Limestone, and gray-green shale, lime mud, and sand stringers  dating back to the Pennsylvanian age.

    The RV Hopkins property is composed of five buildings: the office and warehouse, a warehouse used
to store drums of hazardous waste and a waste pile, a manufacturing building, a drum reclamation furnace,
and a cutting shed. The office and the warehouse are located on the southwest corner of the site. Areas
investigated on each site include the furnace area, the old  and new baghouses, the former drum storage area
on the north end of the facility, the former landfill, and a drainage ditch. Major contaminants include
barium, lead, chromium, and zinc, as well as lesser concentrations of other metals, such as copper and
nickel, pesticides, and volatile organic compounds.

    Based on historical data, the most concentrated contaminants in the  furnace area are chromium, lead,
and zinc. The highest concentrations of these elements are at the furnace entrance, as opposed to the
furnace exit. The concentrations of lead are higher in the  old baghouse than in the new, while the new
baghouse exhibits a higher concentration of chromium, as well  as high iron, lead, and barium
concentrations. The former landfill has concentrations of barium, chromium, lead, nickel, and zinc greater
than 1,000 mg/kg.  Lead is the most prevalent contaminant in the former drum storage area with lesser
concentrations of barium, chromium, and zinc.

Predemonstration Sampling

    Predemonstration sampling was conducted at both sites between December 5 and 14, 1994. These
sampling events had the following objectives:

    • To provide data on, or verify, the extent of surface contamination at each site and to locate optimum
     sampling areas for the demonstration.

    • To allow the developers to analyze samples from the demonstration sites in advance of the
     demonstration, and if necessary, refine and recalibrate their technologies and revise their operating
     instructions.

    • To evaluate samples for the presence of any unanticipated matrix effects or interferences that might
     occur during the demonstration.

    • To check the quality assurance (QA) and QC procedures of the reference laboratory.

    One hundred soil samples were analyzed on each site by the FPXRF analyzers during the
predemonstration sampling activities. The  samples represented a wide range in the concentration of metals
and soil textures.  Thirty-nine samples were submitted for reference method analysis using EPA SW-846
Methods 3050A/6010A. Twenty-nine of these samples were split and sent to the developers. Nine field
duplicates were collected and submitted for reference method analysis to assess proposed sample
homogenization procedures. One purchased PE sample also was submitted to the reference laboratory to
provide an initial check of its accuracy.

-------
    Additionally, three samples representing low, medium, and high concentrations were collected at each
site. These samples were dried, ground, and then analyzed by six independent laboratories before the
demonstration began to create site-specific PE samples. These samples were analyzed with
laboratory-grade X-ray fluorescence (XRF) analyzers.

Experimental Design

    The experimental design of this demonstration was developed to meet the primary and secondary
objectives stated above, and was approved by all demonstration participants prior to the start of the
demonstration.  The design is detailed in the demonstration plan (PRC 1995) and is  summarized below.

    Approximately 100 soil samples were collected from each of three target soil textures: clay, loam, and
sand. This variety of soil textures allowed the examination of the effect of soil texture on data
comparability. Splits of these samples were analyzed by all of the FPXRFs and by the reference methods.

    The X-MET 920-MP Analyzer can be operated in either an in situ or intrusive mode. These two modes
of analysis involve  slightly different measurement and preparation procedures. These procedures allowed
for  an evaluation of the effects of sample preparation or FPXRF comparability to the reference data. For
in situ analysis, an  area 4 inches by 4 inches square was cleared of vegetation, debris, and gravel larger
than 2 millimeters (mm) in diameter.  Each analyzer took one in situ measurement in each sample area.
These data represented FPXRF in situ measurements for unprepared oils (in situ -unprepared).  Replicate
measurements were taken at 4 percent of these locations to assess analyzer precision. Figure 2-1 depicts
the  sample analysis flowchart for both in situ and intrusive analyses.

    After the in s/YH-unprepared analysis was complete at a given location, the soil within the 4-inch by 4-
inch square was removed to a depth of 1 inch and placed in a plastic bag.  This produced a soil sample of
approximately 375  grams or 250 cubic centimeters (cm3).  Sample homogenization was monitored by
adding 1 to 2 grams of sodium fluorescein salt (which fluoresces when exposed to ultraviolet light) to the
sample bag. During the predemonstration, it was determined that sodium  fluorescein did not affect the
FPXRF or reference method analysis. Sample homogenization took place by kneading the sample and
sodium fluorescein salt in the bag for 2 minutes. After this period, the sample preparation technician
examined  the  sample under ultraviolet light to assess the distribution of sodium fluorescein. If the sodium
fluorescein salt was not evenly distributed, the homogenization and checking process were repeated until it
was evenly distributed. This monitoring process assumed that thorough distribution of sodium fluorescein
was indicative of good sample homogenization.  The effectiveness of this process is  discussed later in this
section.

    The homogenized sample was then spread out inside a 1-inch-deep petri dish. The FPXRF analyzer
took one measurement from this homogenized material. This represented the homogenized sample analysis
for the in situ analyzers (in situ -prepared). This approximated the common practice of sample
homogenization in  a plastic bag and subsequent sample measurement through the bag. Replicate
measurements were also collected from 4 percent of these samples to assess analyzer precision. These
replicate measurements were made on the same soils as the unprepared precision measurements.

    Following the in situ -prepared procedure, the sample material was passed through a No. 10 mesh sieve
(2-mm openings) and approximately 10 grams of this material was placed in a sample cup for analysis in
an intrusive mode.  The same sample cup was used for each FPXRF analyzer. Replicate measurements
were collected from 4 percent of these samples to assess analyzer precision. These replicate measurements

                                               8

-------
were made on the same soils as the in s/YM-prepared precision measurements.  These data represented
FPXRF intrusive mode measurements on soils with no sample preparation (intrusive-unprepared).  Sample
material from this preparation step was collected and submitted to the reference laboratory for reference
method analysis.
    Figure 2-1. Sample Preparation and Analysis: This flowchart depicts the handling
    procedures for each sample taken during the demonstration.

-------
    Following the intrusive-unprepared analysis, a portion of the soil sample was dried in a convection
oven at 110 °C for 1 hour and ground with a mortar and pestle until it passed through a No. 40 stainless-
steel sieve (0.425-mm openings).  The sample was then analyzed in the intrusive mode. Four percent of
these samples underwent replicate measurements to evaluate analyzer precision. These replicate
measurements were performed on the same soils as in the intrusive-unprepared precision measurements.
This data represented FPXRF intrusive measurements on prepared soils (intrusive-prepared).

Qualitative Factors

    There were a number of factors important to data collection that are difficult to quantify and must be
evaluated qualitatively.  These are considered qualitative factors.  One such factor was the amount of
training required to operate a given FPXRF analyzer. To assess this factor, PRC operators were trained by
the developers on how to operate  their respective FPXRF analyzers.  All operators met or exceeded the
developers' minimum requirements for education and previous experience. Demonstration procedures were
designed to simulate routine field conditions as closely as possible. Based on this training and field
experience, the operators prepared a subjective evaluation assessing  the training and technology operation
during the demonstration (Section 4).

    Many analytical methods exhibit "operator effects," in which individual differences in sample
preparation or operator technique result in a significant effect on the numerical results. To reduce the
possible influence of operator effects, a single operator was used to operate each FPXRF analyzer. While
this reduced some error from the evaluation, it did not allow the analyzers to be evaluated for their
susceptibility to operator-induced error. A single operator was used for all the FPXRFs demonstrated
except the X-MET 920-MP.  Due to unforeseen circumstances, a second operator was needed to analyze
samples for the RV Hopkins site.  Sample preparation variation effects were minimized by using the  same
personnel to prepare samples.  To eliminate the influence of operator effects on the reference method
analysis, only one reference laboratory was used to analyze the samples.  Based on this design, there  could
be no quantitative estimate of "operator" effect.

Quantitative Factors

    Many factors in this demonstration could be quantified by various means. Examples of quantitative
factors evaluated during this demonstration include analyzer performance near regulatory action levels, the
effects of sample preparation, effects of microwave sample drying, count times, health and safety
considerations, costs, and interferences.

    The data developed by the FPXRF analyzers were compared to reference data for the following
primary analytes:  arsenic, barium, chromium, copper, lead, and zinc; and for the following secondary
analytes: nickel, iron, cadmium, and antimony. The specific target analytes determined by the X-MET
920-MP in this demonstration were arsenic, barium, copper, chromium, lead, nickel, and zinc.

    Evaluations of analyzer data comparability involved examining the effects of each site, soil texture, and
sample preparation technique (Table 2-1). Two sites were sampled for this demonstration and, therefore,
two site variables were examined (RV Hopkins and ASARCO sites). These sites produced samples from
three distinct soil textures and, therefore, three soil variables were examined (clays, sands, and loams).
Four sample preparation steps were used:  (1) in s/YH-unprepared, (2) in s/YH-prepared, (3) intrusive-
unprepared, and (4) intrusive-prepared. These variables were nested as follows: each site was divided into
RV Hopkins and ASARCO data sets; the RV Hopkins data represented the clay soil texture, and the

                                               10

-------
ASARCO data was divided into sand and loam soil textures; each soil texture was subdivided by the soil
preparation steps. This design allowed for the examination of particle size and sample homogenization.
These effects were believed to have the greatest impact on data comparability.

    Of greatest interest to users is analyzer performance near action levels.  For this reason, samples were
approximately distributed as follows:  25 percent in the 0 - 100 mg/kg range, 50 percent in the 100 - 1,000
mg/kg range, and 25 percent in the greater than 1,000 mg/kg range.  The lower range tested analyzer
performance near MDLs; the middle range tested analyzer performance in the range of many action levels
for inorganic contaminants; and the higher range tested analyzer performance on grossly contaminated
soils.  All samples collected for the demonstration were split between the FPXRF  analyzers and reference
laboratory for analysis. Metal concentrations measured using the reference methods were considered to
represent the "true" concentrations in each sample.  Where duplicate samples existed, concentrations for
the duplicates were averaged and the average concentration was considered to represent the true value for
the sample pair.  This procedure was specified in the demonstration plan.  If one or both samples in a
duplicate pair exhibited a nondetect for a particular target analyte, that pair of data was not used in the
statistical evaluation of that analyte. The reference methods reported measurable concentrations of target
analytes in all of the samples analyzed.

                Table 2-1. Performance and Comparability Variables Evaluated
Variables
Site Name (315)
ASARCO (2 15)

RV Hopkins (100)
Soil Texture (31 5)
Sand (100)
Loam (11 5)
Clay (1 00)
Preparation Step [1,260]
in situ-unprepared [100]
in situ-prepared [100]
intrusive-unprepared [100]
intrusive-prepared [100]
in situ-unprepared [115]
in situ-prepared [115]
intrusive-unprepared [115]
intrusive-prepared [115]
in situ-unprepared [100]
in situ-prepared [100]
intrusive-unprepared [100]
intrusive-prepared [100]
                Notes:  (  ) Total number of sample points.
                       [  ] Total number of measurements taken.

    In addition to the quantitative factors discussed above, the common FPXRF sample preparation
technique of microwave drying was evaluated.  Sample temperatures during this procedure can be high
enough to melt some mineral fractions in the sample or combust organic matter. Several metals that
present environmental hazards can volatilize at elevated temperatures.  Arsenic sublimes at 188 °C, within
the temperature range achievable during microwave drying.  To assess this effect, 10 percent of the
homogenized, crushed, oven-dried, and sieved (No. 40) samples were split and heated in a microwave oven
on high for 3 minutes. This time was chosen to approximate the common microwave drying times used in
the field. These split samples were then submitted for reference analysis. The reference data for these
samples were compared to the corresponding reference data produced from the convection oven-dried
sample.  These data showed the effects of the microwave drying variable on analyte concentration. This
was a minor variable and was only evaluated for the reference laboratory in an attempt to identify potential
effects on data comparability.
                                               11

-------
    Another quantitative variable evaluated was the count time used to acquire data. During the formal
sample quantitation and precision measurement phase of the demonstration, the count times were set by the
developers and remained constant throughout the demonstration. Count times can be tailored to prod-
uce the best results for specific target analytes. The developers, however, selected count times that
produced the best compromise of results for the entire suite of target analytes. To allow a preliminary
assessment of the effect of count times, select soil samples were analyzed in replicate using count times
longer and shorter than those set by the developers. This allowed the evaluation of the effects of count
times on analyzer performance.

    An important health and safety issue during the demonstration was the effectiveness of radioactivity
shielding of each FPXRF analyzer. Qualitative radiation readings were made with a gamma ray detector
near each  analyzer to assess the potential for exposure to radiation.

    A compilation of the costs associated with the use of each FPXRF analyzer was another important
evaluation factor. Cost includes analyzer purchase or lease, expendable supplies, such as liquid nitrogen
and sample cups, and nonexpendable costs, such as labor, licensing  agreements for the radioactive sources,
operator training costs, and disposal of investigation-derived waste (IDW).  This information is provided to
assist the user in developing a project cost analysis.

    Factors that could have affected the quantitative evaluations included interference effects and matrix
effects.  Some  of these effects and the procedures used to evaluate their influence during this demonstration
are summarized below:

    • Heterogeneity:  For in s/YH-unprepared measurements, heterogeneity was partially controlled by
      restricting  measurements within a 4-by-4-inch area. For measurements after the initial point-and-
      shoot preparation, heterogeneity was minimized by sample homogenization. This effect was
      evaluated through the sample preparation data.

    • Particle Size: The effect of particle size was evaluated using the two intrusive sample preparation
      procedures.  Theoretically, precision and accuracy should increase as particle size decreases and
      becomes more uniform.

    • Moisture Content: It has been suggested that major shifts in sample moisture content can affect a
      sample's relative fluorescence. This effect could not be evaluated as thoroughly as planned because
      of the small difference in sample moisture content observed at the two sites.

    • Overlapping Spectra of Elements: Interferences result from overlapping spectra of metals that  emit
      X-rays with similar energy levels.  The reference method analysis provided data on the concentration
      of potential interferants in each sample.

Evaluation of Analyzer Performance

    Metals concentrations measured by each analyzer were compared to the corresponding reference
laboratory data and to other QA/QC sample results. These comparisons were conducted independently for
each target analyte. These measurements were used to determine an analyzer's accuracy, data quality
level, method precision, and comparability to reference methods. PE samples and SRM samples were used
to assess analyzer accuracy. Relative standard deviations (RSD) on replicate measurements were used to
determine analyzer precision.  These data were also used to help determine the data quality of each FPXRF

                                               12

-------
analyzer's output. The data comparability and quality determination was primarily based on a comparison
of the analyzer's data and the reference data.  Linear regression and a matched pairs t-test were the
statistical tools used to assess comparability and data quality.

    A principal goal of this demonstration was the comparison of FPXRF data and the reference laboratory
data. EPA SW-846 Methods 3050A/6010A were selected as the reference methods because they represent
the  regulatory standard against which FPXRF is generally compared. In comparing the FPXRF data and
reference data, it is important to recognize that,  while similar, the process from which the data are obtained
is not identical.  While there  is  significant overlap in the nature of the analysis, there are also major
differences. These differences, or "perspectives," allow the user to characterize the same sample in slightly
different ways.  Both methods have a role in site characterization and remediation.  It is important to
consider these differences and the measurement error intrinsic to each method when comparing the FPXRF
method against a reference analytical method.

    The reference laboratory methods involve wet chemical analysis and partial acid digestion of
approximately 1 to 2 grams of sample (approximately 0.25 cubic centimeters (cm3) depending on sample
bulk density). The digestion process extracts the most acid-soluble portion of the sample. Since the
digestion is not complete, the less acid-soluble components are not digested and will not be included in the
analysis. These components may include the coarser-grained quartz, feldspar, lithic components, and
certain metal complexes. In  contrast, FPXRF analyzers generally produce X-ray excitation in an area of
approximately 3 cm2 to a depth of approximately 2.5 centimeters  (cm).  This equates to a sample volume of
approximately 7.5 cm3.  X-rays returning to the detector are derived from all matrix material including the
larger-grained quartz, feldspar, lithic minerals, metal complexes, and organics. Because the FPXRF
method analyzes all material, it represents a total element analysis in contrast to the reference methods,
which may only represent a select or partial analysis. This difference can  result in FPXRF concentrations
that are higher than the corresponding reference data when metals are contained within a nonacid  soluble
complex.  It is important to note that if metals are contained in nonacid soluble complexes, a difference
between the FPXRF analyzers and the reference methods is not necessarily due to error in the FPXRF
result but rather to inherent differences in the sample preparation procedures.

    The comparison of FPXRF data and the reference data employs a linear regression as the primary
statistical tool. Linear regression analysis intrinsically contains assumptions and conditions that must be
valid for each data set.  Three important assumptions are: (1) the linearity of the relationship, (2) the
confidence interval and constant error variance, and (3) an insignificant measurement error for the
independent variable (reference data).

    The first assumption requires that the independent variable (reference data) and the dependent variable
(FPXRF data) are linearly related and are not described by some curvilinear or more complex relationship.
This linearity condition applies to either the raw data or mathematical transformations of the raw data.
Figure 2-2 illustrates that the FPXRF data and the reference data are, in fact, related linearly and that this
assumption is correct.

    The second assumption requires that the error be normally distributed, the sum to equal zero, be
independent, and exhibit a constant error variance for the data set. Figure 2-2 illustrates that for raw data,
this assumption is not correct (at higher concentrations the scatter around the regression line increases), but
that for the logarithmic transformation (shown as a log-log plot) of the data, this assumption is valid (the
scatter around the regression line is uniform over the entire concentration range). The change in error
distribution (scatter) evident  in  the untransformed data results in the disproportionate influence of large
data values compared with small data values on the  regression analysis.

                                                13

-------
                  Linear Data Plot - Lead
                         Thousands
                   Reference Data (mg/kg)
                                                       Linear Data Plot - Arsenic
                                                         10

                                                          8
                                              in
                                              T3
                                           II  6
                                           O =>  A
                                           CM °  4
                                           en £
                                                    LU
                                                                                           n

                                                                                            n
                                                         246!
                                                               Thousands
                                                         Reference Data (mg/kg)
                                                                                             10
                Log-Log Data Plot - Lead
                                                      Log-Log Data Plot ~ Arsenic
       10000^
    O)
    CD
    CM
    O>

    LU
        1000 r
                                              10000
100
£

Q_

CD
CM
O)

LLJ
                                               1000  -
                      100        1000
                   Reference Data (mg/kg)
                                   10000
100
             100        1000
          Reference Data (mg/kg)
                                       10000
  Figure 2-2. Linear Log-log Data Plots:  These graphs illustrate the linear relationship between the
  FPXRF data and the reference data. The linear data plots illustrate the concentration dependence of
  this relationship with the increased scatter at higher concentrations. The log-log plots eliminate this
  concentration effect.  Scatter is relatively constant over the entire plot.
    The use of least squares linear regression has certain limitations. Least squares regression provides a
linear equation, which minimizes the squares of the differences between the dependent variable and the
regression line. For data sets produced in this demonstration, the variance was proportional to the
magnitude of the measurements.  That is, a measurement of 100 parts per million (ppm) may exhibit a 10
percent variance of 10 ppm, while a 1,000 ppm measurement exhibits a 10 percent variance of 100 ppm.
For data sets with a large range in values, the largest measurements in  a data set exert disproportionate
influence on the regression analysis because the least squares regression must account for the variance
associated with the higher valued measurements. This can result in an  equation that has minimized error
for high values, but almost neglects error for low values because their influence in minimizing dependent
variable error is small or negligible.  In some cases, the resulting equations, biased by high-value data, may
lead to inappropriate conclusions concerning data quality.  The range of the data examined for the
analyzers spanned between 1 and 5 orders of magnitude (e.g., 10 - 100,000 ppm) for the target analytes.
This wide range in values and the associated wide range in variance (influenced by concentration) created
the potential for this problem to occur in the demonstration data set. To provide a correlation that was
equally influenced by both high and low values, logarithms (Iog10) of the dependent and independent
variables were used, thus, scaling the concentration measurements and  providing equal weight in the  least
squares regression analysis to both small and large values (Figure 2-2). All statistical evaluations were
carried out on Iog10 transformed data.
                                                14

-------
    The third assumption, requiring an insignificant measurement error in the reference data, was not true
for all analytes. The consequences of measurement error varied depending on whether the error is caused
by the reference methods or the FPXRF method. If the error is random or if the error for the reference
methods is small compared to the total regression error, then conventional regression analysis can be
performed and the error becomes a part of the random error term of the regression model. This error
(based on the Iog10 transformed data) is shown in the regression summary tables in Section 4 as the
"standard error."  In this case, deviations from perfect comparability can be tied to an analyzer's
performance. If the error for the reference methods is large compared to the total error for the correlation
of the FPXRF and the reference data, then deviations from perfect comparability might be due in part to
measurement error in the reference methods.

    It is a reasonable assumption that any measurement errors in either the reference or FPXRF methods
are independent of each other.  This assumption applies to either the raw data or the  Iog10 transformed data.
Given this assumption, the total regression error is approximately the sum of the measurement error
associated with the reference methods and the measurement error associated with the FPXRF method. The
reference methods' precision is a measure of independent variable error, and the mean square error
expressed in the regression analysis is a relative measure of the total regression error that was determined
during the regression analysis.  Precision data for the reference methods, obtained from RPD analyses on
the duplicate samples from each site, for each analyte, indicated the error for the reference methods  was
less than 10 percent of the total regression error for the target analytes.  Subsequently, 90 percent of the
total measurement error can be attributed to measurement error associated with the analyzers.

    The comparison of the reference data to the FPXRF data is referred to as an intermethod comparison.
All reference and QA/QC data were generated using an EPA-approved definitive level analytical method.
If the data obtained by an analyzer were statistically similar to the reference methods, the analyzer was
considered capable of producing definitive level data.  As the statistical significance  of the comparability
decreased, an analyzer was considered to produce data of a correspondingly lower quality. Table 2-2
defines the criteria that determined the  analyzer's level of data quality (EPA 1993).

    Results from this demonstration were used to assign analyzer data into one of three data quality levels
as follows: (1) definitive, (2) quantitative screening, and (3) qualitative screening. The first two data
quality levels are defined in EPA guidance (1993).  The qualitative screening level criteria were defined in
the demonstration plan (PRC 1995) to further differentiate screening level data.

    Definitive level data are considered the highest level of quality.  These data are usually generated by
using rigorous, well-defined analytical methods such as approved EPA or ASTM methods.  The data is
analyte-specific with full confirmation  of analyte identity and concentration. In addition, either analytical
or total measurement error must be determined. Definitive data may be generated in the  field as long as the
project QA/QC requirements are  satisfied.

    Quantitative screening data provides confirmed analyte identification and quantification, although the
quantification may be  relatively imprecise.  It is commonly recommended that at least  10 percent of
screening level data be confirmed using the appropriate analytical methods and QA/QC procedures  and
criteria  associated with definitive data.  The quality of unconfirmed screening data cannot be determined.

    Qualitative screening level data indicates the presence or absence of contaminants in a sample matrix,
but does not provide reliable concentration estimates.  The data may be compound-specific or specific to
                                                15

-------
classes of contaminants. Generally, confirmatory sampling is not required if an analyzer's operation is
verified with one or more check samples.
  Table 2-2.  Criteria for Characterizing Data Quality
      Data Quality Level
  Definitive Level
                      Statistical Parameter3'13
r2 = 0.85 to 1.0. The precision (RSD) must be less than or equal to 10
percent and the inferential statistics must indicate that the two data
sets are statistically similar.
  Quantitative Screening
  Level
r2 = 0.70 to 1.0. The precision (RSD) must be between 10 and 20
percent, but the inferential statistics indicate that the data sets are
statistically different.
  Qualitative Screening
r2 = less than 0.70. The precision (RSD) is greater than 20 percent.
The data must have less than a 10 percent false negative rate.
  Notes:         The statistical tests and parameters are discussed later in the "Intermethod Comparison"
                subsection in Section 4.
              b
                The regression parameters apply to either raw or Iog10 transformed data sets. The precision
                criteria apply to only the raw data.
             r2  Coefficient of determination.
           RSD  Relative standard deviation.

    At the time of this demonstration, approved EPA methods for FPXRF did not exist. As part of this
demonstration, PRC prepared draft Method 6200 "Field Portable X-Ray Fluorescence Spectrometry for the
Determination of Elemental Concentrations in Soil and Sediment." The draft method has been submitted
for inclusion in Update 4 of SW-846 scheduled for approval in FY-97. For purposes of this demonstration,
the absence of an EPA-approved final method did not preclude the analyzers' data from being considered
definitive. The main criterion for data quality level determination was based on the comparability of each
analyzer's data to that produced by the reference methods, as well as analyzer-specific criteria such as
precision.

    The comparability data set for the X-MET 920-MP Analyzer consisted of matched pairs of FPXRF
and reference data for each target analyte.  Each data set was analyzed as a whole and then subdivided and
analyzed with respect to each of the variables listed in Table 2-1.  This nesting of variables allowed for an
independent assessment of the influence of each variable on comparability.

    For the purposes of this demonstration, a total of 315  soil samples was collected for analysis.  These
samples were analyzed by the reference methods and by the FPXRF analyzer up to four times, using each
of the four sample preparation steps. This would produce 1,260 data values for these  analyzers which used
all four of the sample preparation steps.  Seventy of the  315 samples submitted to the  reference laboratory
were split and submitted as field duplicates to assess the sample homogenization process. Thirty-three of
the 315 samples were also split and microwave-dried; then submitted for reference method analysis to
assess the effect of microwave drying. Of the 315 samples submitted for reference method analysis, 215
were collected from the ASARCO site and 100 were collected from the RV Hopkins site. Approximately
twice as many samples were  collected at the ASARCO  site because two of the target soil textures (sands
and loams) were found there. Only one target soil texture (clay) was  found at the RV  Hopkins site.

    Evaluation of the influence of the site and soil variables was limited to an examination of the lead and
zinc data. These were the only primary analytes that exhibited a wide distribution of concentrations across
                                                16

-------
all sites and soil textures. The effects of sample preparation were evaluated for all target analytes. If the
evaluation of the influence of a given variable did not result in a better correlation, as exhibited by a higher
coefficient of determination (r2) and smaller standard error of the estimate (using Iog10 transformed data),
then the influence was considered to be insignificant.  However, if the correlation worsened, the cause was
examined and explained. If the correlation improved, resulting in a higher r2 and reduced standard error of
the estimate, then the impact of the variable was considered significant. For example, if the r2 and standard
error of the estimate for a given target analyte improved when the data set was divided into the four sample
preparation steps, the sample preparation variable was determined to be significant.  Once this was
determined, the variables of site and soil texture were evaluated for each of the four sample preparations
steps. If the site or soil texture variable improved the regression parameters for a given soil preparation,
then that variable was also considered significant.

    After the significant variables were identified, the impact of analyte concentration was examined. This
was accomplished by dividing each variable's Iog10 transformed data set into three concentration ranges:  0
- 100 mg/kg; 100 - 1,000 mg/kg; and greater than 1,000 mg/kg.  A linear regression  analysis was then
conducted on these data sets.  If this did not result in improved r2 values and reduced standard errors  of the
estimate, then the relationship between the analyzer's Iog10 transformed data and the  Iog10 transformed
reference data was considered linear over the entire range of concentrations encountered during the
demonstration.  This would mean that there was no concentration effect.

    Numerous statistical tests have  been designed to evaluate the significance of differences between two
populations. In comparing the performance of the FPXRF analyzers against the reference methods, the
linear regression comparison and the paired t-test were considered the optimal statistical tests. The paired
t-test provides a classic test for comparing two populations, but is limited to analysis of the average or
mean difference between those populations. Linear regression analysis provides information not only about
how the two populations compare on average, but also about how they  compare over ranges of values.
This statistical analysis technique provides information about the structure of the relationship; that is,
whether the methods differ at high  or low concentrations or both. It also indicates whether the FPXRF data
is biased or shifted relative to the reference data.

    Linear regression provides an equation that represents a line (Equation 2-1). Five linear regression
parameters were considered when assessing the level of data quality produced by the FPXRF analyzers.
This assessment was made on the Iog10 transformed data sets.  The five parameters were the y-intercept, the
slope of the regression line, standard error of the estimate, the correlation coefficient (r), and r2. In linear
regression analysis, the r provides a measure of the degree or strength of the correlation between the
dependent variable (Iog10 transformed FPXRF data), and the independent variable  (Iog10 transformed
reference data). The r2 provides a measure of the fraction of total variation which is accounted for by the
regression relation (Havlick and Grain 1988). That is, it is  a measure of the scatter about a regression line
and, thus, is a measure of the strength of the linear association.

      Y =  mX +  b                                                                             (2-1)

 where

      b is  the y-intercept of the  regression line,  m  is the  slope of the regression line,
      and Y and X are the  logw transformed dependent and independent variables, respectively.

    Values for r vary from 1 to -1,  with either extreme indicating a perfect positive or negative correlation
between the independent and dependent variables.  A positive correlation coefficient  indicates that as  the

                                                17

-------
independent variable increases, the dependent variable also increases.  A negative correlation coefficient
indicates an inverse relationship, as the independent variable increases the dependent variable decreases.
An r2 of 1.0 indicates that the linear equation explains all the variation between the data sets. As the r2
departs from 1.0 and approaches zero, there is more unexplained variation, due to such influences as lack
of association with the dependent variable (Iog10 transformed FPXRF data), or the influence of other
independent variables.

    If the regression correlation exhibited an r2 between 0.85 and 1.0, the FPXRF data was considered to
have met the first requirement for definitive level data classification (Table 2-2).  The second criteria,
precision is then examined and required to be equal to or less than 10 percent RSD to maintain the
definitive data quality level designation.  If both these criteria are not satisfied, then additional inferential
statistical parameters are evaluated. The regression line's y-intercept and slope would be examined next. A
slope of 1.0 and a y-intercept of 0.0 would mean that the results of the FPXRF analyzer matched those of
the reference laboratory (Iog10 FPXRF=log10 reference). Theoretically, the more  the slope and y-intercept
differ from the values of 1.0 and 0.0, respectively, the less accurate the FPXRF analyzer. However, a
slope or y-intercept can differ slightly from these values without that difference being statistically
significant.  To determine whether such differences were statistically significant, the Z test statistics for
parallelism and for a common intercept was used at the 95 percent confidence level for the comparison
(Equations 2-2 and 2-3) (Kleinbaum and Kupper 1978).  This process was used to assign data quality
levels for each reported analyte.

      Slope Test for Significant Differences                                                     (2-2)

              m - 1
     Z  =
where
      m is the  slope  of the regression line, SE is the standard error  of the slope,
      and Z is the normal deviate test statistic.
      Y-intercept Test for Significant Differences                                               (2-3)

      Z  =  	
where

      b  is the y-intercept of the regression line, SE is  the standard error of the slope,
      and Z is the normal deviate test statistic.


    The matched pairs t-test was also used to evaluate whether the two sets of Iog10 transformed data were
significantly different.  The paired t-test compares data sets, which are composed of matched pairs of data.
The significance of the relationship between two matched-pairs sets of data can be determined by
comparing the calculated t-statistic with the critical t-value determined from a standard t-distribution table
at the desired level of significance and degrees of freedom.  To meet definitive level data quality
requirements, both the slope and y-intercept had to be statistically the same as their ideal values, as defined
in the demonstration plan, and the data had to be statistically similar as measured by the t-test.  Log10

                                                18

-------
transformed data meeting these criteria were considered statistically equivalent to the Iog10 transformed
reference data.

    If the r2 was between 0.70 and 1, the precision was less than 20 percent RSD, and the slope or intercept
were not statistically equivalent to ideal values, the analyzer was considered to produce quantitative
screening level data quality. However, the linear regression was deemed sufficiently significant that bias
could be identified and corrected. Results in this case could be mathematically corrected if 10 - 20 percent
of the samples are sent to a reference laboratory.  Reference laboratory analysis results for a percentage of
these samples would provide a basis for determining a correction factor.

    Data placed in the qualitative screening level category exhibit r2 values less than 0.70. These data
either were not statistically similar to the reference data based on inferential statistics or had a precision of
greater than 20 percent RSD. An analyzer producing data at this level is considered capable of detecting
the presence or lack of contamination, above its detection limit, with at least a 90 percent accuracy rate, but
is not considered suitable for reporting of concentrations.

    MDLs for the analyzers were determined in two ways.  One approach followed standard SW-846
protocol. In this approach, standard deviations (SD) from precision measurements of samples exhibiting
contamination 5 to 10 times the estimated detection levels of the analyzers were multiplied by 3. The result
represents the MDL for the analyzer.

    In a second approach, MDLs were determined by analysis of the low concentration outliers on the Iog10
transformed FPXRF and Iog10 transformed reference method data cross plots.  These cross plots for all
analytes characteristically exhibited a region below the  MDL where the linearity of the relationship
disintegrated. Above the  MDL, the FPXRF concentrations increased linearly with increasing reference
method values.  Effectively, the linear correlation between the two methods abruptly changes to no
correlation at a point below the MDL.  By determining  the point where the linear relationship disintegrates,
the MDL is  assigned at two SDs above this concentration. This procedure represents a field- or
performance-based MDL.

Deviations from the Demonstration Plan

    Seven deviations were made from the demonstration plan during on-site activities. The first dealt with
the determination of the moisture content of the samples.  The demonstration plan stated that a portion of
the original sample would be used for determining moisture content.  Instead, a small portion of soil
immediately adjacent to the original  sample location was used for determining moisture content. This was
done to conserve sample volume for the reference laboratory.  The moisture content sample was not put
through the homogenizing and sieving steps prior to drying.

    The second deviation dealt with the sample drying procedures for moisture content determination. The
demonstration plan required that the  moisture content samples be dried in a convection oven at   150 °C
for 2 hours.  Through visual observation, it was found that the samples were completely dried in 1 hour
with samples heated to only 110 °C. Therefore, to conserve time, and to reduce the potential volatilization
of metals, the samples for moisture content determination were dried in a convection oven at 110 °C for 1
hour.

    The third deviation involved an assessment of analyzer drift due to changes in temperature. The
demonstration plan required that at each site, each analyzer  would measure the same SRM or PE sample at

                                                19

-------
2-hour intervals during at least one day of field operation.  However, since ambient air temperature did not
fluctuate more than 20 °F on any day throughout the demonstration, potential analyzer drift due to changes
in temperature was not assessed.

    The fourth deviation involved the drying of samples with a microwave. Instead of microwaving the
samples on high for 5 minutes, as described in the demonstration plan, the samples were microwaved on
high for only 3 minutes.  This modification was made because the plastic weigh boats, which contained the
samples, were melting and burning when left in the microwave for 5 minutes. In addition, many of the
samples were melting to form a slag. PRC found (through visual observation) that the samples were
completely dry after only 3 minutes.  This interval is within common microwave drying times used in the
field.

    An analysis of the microwaved samples showed that this process had a significant impact on the
analytical results.  The mean RPD for the microwaved and nonmicrowaved raw data were significantly
different at a 95 percent confidence level. This suggests that the microwave drying process somehow
increases error and sample concentration variability.  This difference may be due to the extreme heat that
altered the reference methods' extraction efficiency for target analytes. For the evaluation of the effects of
microwave drying, there were 736 matched pairs of data where both element measurements were positive.
Of these pairs, 471 exhibited RPDs less than 10 percent. This 10 percent level is within the acceptable
precision limits for the reference laboratory as defined in the demonstration QAPP. Pairs exhibiting RPDs
greater than 10 percent totaled 265. RPDs greater than 10 percent may have causes other than analysis-
induced error.  Of these 265, 96 pairs indicated an increase in metals concentration with microwaving, and
169 pairs indicated a reduction in metals concentration. The RPDs for the microwaved samples were 2 to
3 times worse than the RPDs from  the field duplicates. This further supports the hypothesis that
microwave drying increases variability.

    The fifth deviation involved reducing the percentage of analyzer precision measuring points.  The
demonstration plan called for 10 percent of the samples to  be used for assessment of analyzer precision.
Due to the time required to complete analysis of an analyzer precision sample, only 4  percent of the
samples were used to assess analyzer precision. This reduction in samples was  approved by the EPA
technical advisor and the PRC field demonstration team leader.  This eliminated 720 precision
measurements and saved up to 3 days of analysis time.  The final precision determinations for this
demonstration were based on 48 sets of 10 replicate measurements for each analyzer.

    The sixth deviation involved method blanks. Method blanks were to be analyzed  each day and were to
consist of a lithium carbonate that had been used in all sample preparation steps. Each analyzer had its
own method blank samples, provided by the developer. Therefore, at the ASARCO site, each analyzer
used its own method blank samples.  However, at the RV Hopkins site, each analyzer used lithium
carbonate method blanks that were  prepared in the field, in addition to its own method blank samples.  Both
types of method blank analysis never identified method-induced contamination.

    The seventh deviation involved assessing the accuracy of each analyzer. Accuracy was to be  assessed
through FPXRF analysis of 10 to 12 SRM or PE samples. Each analyzer measured a total of 28  SRM or
PE samples. Instead, PE samples were used to evaluate the accuracy of the reference methods, and SRMs
were used to evaluate the accuracy of the analyzers.  This is because the PE concentrations are based on
acid extractable concentrations while SRM concentrations represent total metals concentration. SRM data
was used for comparative purposes for the reference methods as were PE data for the FPXRF data.
                                               20

-------
Sample Homogenization

    A key quality issue in this demonstration was ensuring that environmental samples analyzed by the
reference laboratory and by each of the FPXRF analyzers were splits from a homogenous sample. To
address this issue, sample preparation technicians exercised particular care throughout the field work to
ensure that samples were thoroughly homogenized before they were split for analysis. Homogenization was
conducted by kneading the soil in a plastic bag for a minimum of 2 minutes. If after this time the samples
did not appear to be well homogenized, they were kneaded for an additional 2 minutes. This continued until
the samples appeared to be well homogenized.

    Sodium fluorescein was used as an indicator of complete sample homogenization. Approximately one-
quarter teaspoon of dry sodium fluorescein salt was added to each  sample prior to homogenization. After
mixing, the sample was examined under an ultraviolet light to assess the distribution of sodium fluorescein
throughout the sample. If the fluorescent dye was evenly dispersed in the sample, homogenization was
considered complete. If the dye was not evenly distributed, the  mixing was continued and checked until the
dye was evenly distributed throughout the sample.

    To evaluate the homogenization process used in this demonstration, 70 field duplicate sample pairs
were analyzed by the reference laboratory. Sample homogenization was critical to this demonstration; it
assured that the  samples measured by the analyzers were as close as possible to samples analyzed by the
reference laboratory. This was essential to the primary objectives  of this demonstration, the evaluation of
comparability between analyzer results and those of the reference methods.

    The homogenization process was evaluated by determining the RPD between paired field duplicate
samples.  The RPDs for the field duplicate samples reflect the total error for the homogenization process
and the analytical method combined (Equation 2-4). When total error was determined for the entire data
set, the resultant mean RPD total (error) and 95 percent confidence interval was 9.7 ± 1.4, for all metals
reported. When only the primary analytes were considered, the RPD total (error) and 95 percent
confidence interval was 7.6 ± 1.2, including the secondary analytes in the RPD calculation which produced
a mean RPD total (error) and a 95 percent confidence interval of 9.3 ± 1.6.
Total Measurement Error  = \j[(Sample Homogenization Error)2 + (Laboratory Error)2]             (2-4)

    Using internal QA/QC data from 27 analyses, it was possible to determine the reference laboratory's
method error. The reference analytical method precision, as measured by the 95 percent confidence interval
around the mean RPDs (laboratory error) of predigestion duplicate analyses, was 9.3 ± 2.9 for all of the
target analytes.

    To determine the error introduced by the sample homogenization alone, the error estimate for the
reference methods was subtracted from the total error (Equation 2-5). Based on the data presented above,
the laboratory-induced error was less than or approximately equal to the total error. This indicates that the
sample homogenization (preparation) process contributed little or no error to the overall sample analysis
process.
Sample Homogenization Error = \/[(Total Measurement Error)2 - (Laboratory Error)2]           (2-5)

    Although the possibility for poorly homogenized samples exists under any homogenization routine, at
the scale of analysis used by this demonstration, the samples were considered to be completely
homogenized.

                                               21

-------
                                         Section 3
                            Reference Laboratory Results
    All soil samples collected from the ASARCO and RV Hopkins sites were submitted to the reference
laboratory for trace metals analysis.  The results are discussed in this section.

Reference Laboratory Methods

    Samples collected during this demonstration were homogenized and split for extraction using EPA SW-
846 Method 3050A. This is an acid digestion procedure where 1 to 2 grams of soil are digested on a hot
plate with nitric acid, followed by hydrogen peroxide, and then refluxed with hydrochloric acid.  One gram
of soil was used for extraction of the demonstration samples.  The  final digestion volume was 100
milliliters (mL). The soil sample extracts were analyzed by Method  6010A.

    Method 6010A provides analysis of metals using Inductively Coupled Plasma-Atomic Emission
Spectroscopy (ICP-AES). This method requires that a plasma be produced by applying a radio-frequency
field to a quartz tube wrapped by a coil or solenoid through which argon gas is flowing. The radio-
frequency field creates a changing  magnetic field in the flowing gas inside the coil, inducing a circulating
eddy current on the argon gas that, in turn, heats it. Plasma is initiated by an ignition source and quickly
stabilizes with a core temperature of 9,000 - 10,000 degrees Kelvin.

    Soil sample extracts are nebulized, and the aerosol is injected into the plasma.  Individual analytes
introduced into the plasma absorb energy and are excited to higher energy states. These higher energy
states have short lifetimes and the individual elements quickly fall  back to their ground energy state by
releasing a photon.  The energy of the emitted photon is defined by the wavelength of electromagnetic
radiation produced. Since many electronic transitions are possible for each individual element, several
discrete emissions at different wavelengths are observed.  Method 6010A provides one recommended
wavelength to monitor for each analyte. Due to complex spectra with similar wavelengths from  different
elements in environmental samples, Method 6010A requires that interference corrections be applied for
quantification of individual analytes.

    Normal turnaround times for the analysis of soil samples by EPA SW-846 Methods 3050A/6010A
range from 21 to 90 days depending on the complexity of the soil samples and the amount of QC
documentation required.  Faster turnaround times  of 1 -  14 days can be obtained, but at additional cost.

    Costs for the analysis of soil samples by EPA  SW-846 Methods 3050A/6010A range from $150 to
$350 per sample depending on turnaround times and the amount of QC documentation required.  A sample
turnaround of 28 days, a cost of $150 per sample, and a CLP documentation report for QC were chosen for
this demonstration.

                                              22

-------
Reference Laboratory Quality Control
    The reference laboratory, Midwest Research Institute (Kansas City, MO), holds certifications for
performing target analyte list metals analysis with the U.S. Army Corps of Engineers-Missouri River
Division, the State of California, and the State of Utah. These certifications include on-site laboratory
audits, data package review audits, and the analysis of PE samples supplied by the certifying agency. PE
samples are supplied at least once per year from each of the certifying agencies.  The reference laboratory's
results for the PE samples are compared to true value results and certifying agency acceptance limits for
the PE samples.  Continuation of these certifications hinges upon acceptable results for the audits and the
PE samples.

    The analysis of soil samples by the reference laboratory was governed by the QC criteria in its SOPs,
Method 6010A, and the demonstration QAPP. Table 3-1 provides QAPP QC requirements that were
monitored and evaluated for the target analytes. Method 6010A QC guidelines also are included in Table
3-1.  Due to the complex spectra derived from the analysis of the demonstration samples, the QAPP QC
requirements were applied only to the primary analytes.  The QAPP QC requirements also were monitored
and evaluated for the secondary analytes and other analytes reported by the reference laboratory. However,
corrective actions were not required  for the secondary analytes.

 Table 3-1.  Reference Laboratory Quality  Control Parameters3
Reference Method
Parameter Frequency Requirement QAPP Requirement
Initial Calibration
Verification (ICV)
Standard
Continuing Calibration
Verification (CCV)
Standard
Initial and Continuing
Calibration Blanks
(ICB)and(CCB)
Interference Check
Standard (ICS)
High Level Calibration
Check Standard
Method Blanks
Laboratory Control
Samples
Predigestion Matrix
Spike Samples
Postdigestion Matrix
Spike Samples
With each initial
calibration
After analysis of every 10
samples and at the end
of analytical run
With each continuing
calibration, after analysis
of every 10 samples, and
at the end of analytical
run
With every initial
calibration and after
analysis of 20 samples
With every initial
calibration
With each batch of
samples of a similar
matrix
With each batch of
samples of a similar
matrix
With each batch of
samples of a similar
matrix
With each batch of
samples of a similar
matrix
±10 percent of true value
±10 percent of true value
±3 standard deviations of
the analyzer background
mean
±20 percent of true value
±5 percent of true value
No QC requirement
specified
No QC requirement
specified
80-120 percent recovery
75-125 percent recovery
±10 percent of true value
±10 percent of true value
No target analytes at
concentrations greater than
2 times the lower reporting
limit (LRL)
±20 percent of true value
±10 percent of true value
No target analytes at
concentrations greater than
2 times the LRL
80-120 percent recovery
80-120 percent recovery
80-120 percent recovery
                                               23

-------
Table 3-1.  Continued
Reference Method
Parameter Frequency Requirement QAPP Requirement
Performance
Evaluation Samples
Predigestion Laboratory
Duplicate Samples
Postdigestion
Laboratory Duplicate
Samples
As submitted during
demonstration
With each batch of
samples of a similar
matrix
With each batch of
samples of a similar
matrix
No QC requirement
specified
20 percent relative
percent difference (RPD)b
No QC requirement
specified
80-120 percent recovery
within performance
acceptance limits (PAL)
20 percent RPDC
10 percent RPDC
 Notes:      Quality control parameters were evaluated on the raw reference data.
           b
            RPD control limits only pertain to original and laboratory duplicate sample results that were greater
            than 10 times the instrument detection limit (IDL).
            RPD control limits only pertain to original and laboratory duplicate sample results that were greater
            than or equal to 10 times the LRL.

    PRC performed three on-site audits of the reference laboratory during the analysis of predemonstration
and demonstration samples.  These audits were conducted to observe and evaluate the procedures used by
the reference laboratory and to ensure that these procedures adhered to the QAPP QC requirements.  Audit
findings revealed that the reference laboratory followed the QAPP QC requirements. It was determined
that the reference laboratory had problems meeting two of the QAPP QC requirements: method blank
results and the high level calibration check standard's percent recovery. Due to these problems, these two
QAPP QC requirements were widened.  The QC requirement for method blank sample results was changed
from no target analytes at concentrations greater than the lower reporting limit (LRL) to two times the
LRL.  The QC requirement for the high level calibration standard percent recovery was changed from ±5 to
±10 percent of the true value. These changes were approved by the EPA and did not affect the results of
the demonstration.

    The reference laboratory internally reviewed its data before releasing it. PRC conducted a QC review
on the data based on the QAPP QC requirements and corrective actions listed in the demonstration plan.

Quality Control Review of Reference Laboratory Data

    The QC data review focused upon the compliance of the data with the QC requirements specified in the
demonstration QAPP. The following sections discuss results from the QC review of the reference
laboratory data.  All QC data evaluations were based on raw data.

Reference Laboratory Sample Receipt, Handling, and Storage Procedures

    Demonstration samples were divided into batches of no more than 20 samples per batch prior to
delivery to the reference laboratory. A total of 23 batches containing 315 samples and 70 field duplicate
samples was submitted to the reference laboratory. The samples were shipped in sealed coolers at ambient
temperature under a chain of custody.

    Upon receipt of the demonstration samples,  the reference  laboratory assigned each sample a unique
number and logged each into its laboratory tracking system.  The samples were then transferred to the
reference laboratory's sample storage refrigerators to await sample extraction.
                                              24

-------
    Samples were transferred to the extraction section of laboratory under an internal chain of custody.
Upon completion of extraction, the remaining samples were returned to the sample storage refrigerators.
Soil sample extracts were refrigerated in the extraction laboratory while awaiting sample analysis.

Sample Holding Times

    The maximum allowable holding time from the date of sample collection to the date of extraction and
analysis using EPA SW-846 Methods 3050A/6010A is 180 days. Maximum holding times were not
exceeded for any samples during this demonstration.

Initial and Continuing Calibrations

    Prior to sample analysis, initial calibrations (ICAL) were performed. ICALs for Method 6010A
consist of the analysis of three concentrations of each target analyte and a calibration blank. The low
concentration standard is the concentration used to verify the LRL of the method.  The remaining standards
are used to define the linear range of the ICP-AES.  The ICAL is used to establish calibration curves for
each target analyte. Method 6010A requires an initial calibration verification (ICV) standard to be
analyzed with each ICAL. The method control limit for the ICV is ±10 percent. An interference check
sample (ICS) and a high level calibration check standard is required to be analyzed with every ICAL to
assess the accuracy of the ICAL. The control limits for the ICS and high level calibration check standard
were ±20 percent recovery and ±10 percent of the true value, respectively.  All ICALs, ICVs, and ICSs met
the respective QC requirements for all target analytes.

    Continuing calibration verification (CCV) standards and continuing calibration blanks (CCB) were
analyzed following the analysis of every 10 samples and at the end of an analytical run. Analysis of the
ICS was also required after every group of 20 sample analyses.  These QC  samples were analyzed to check
the validity of the ICAL. The control limits for the CCVs were ±10 percent of the true value.  The control
limits for CCBs were no target analyte detected at concentrations greater than 2 times the LRL. All CCVs,
CCBs, and ICSs  met the QAPP requirements for the target analytes with the exception of one CCV where
the barium recovery was outside the control limit. Since barium was a primary analyte, the sample batch
associated with this CCV was reanalyzed and the resultant barium recovery met the QC criteria.

Detection Limits

    The reference laboratory LRLs for the target analytes are listed in Table 3-2. These LRLs were
generated through the use of an MDL study of a clean soil matrix. This clean soil matrix was also used for
method blank samples and LCSs during the analysis of demonstration samples.  The MDL study involved
seven analyses of the clean soil matrix spiked with low concentrations of the target analytes. The mean and
standard deviation of the response for each target analyte was calculated. The LRL was defined as the
mean plus three times the standard deviation of the response for each target analyte included in the method
detection limit study. All LRLs listed in Table 3-2 were met and maintained throughout the analysis of the
demonstration samples.

    The reference laboratory reported soil sample results in units of milligram per kilogram wet weight.
All reference laboratory results referred to in this report are wet-weight sample results.
                                              25

-------
                      Table 3-2.  SW-846 Method 601OA LRLs for Target
                                 Analytes
Analyte LRL (mg/kg) Analyte LRL (mg/kg)
Antimony
Arsenic*
Barium*
Cadmium
Chromium*
6.4
10.6
5.0
0.8
2.0
Copper*
Iron
Lead*
Nickel
Zinc*
1.2
600a
8.4
3.0
2.0
                      Notes:
      LRL elevated due to background
      interference.
    * Primary analyte.
mg/kg Milligrams per kilogram.
Method Blank Samples
    Method blanks were prepared using a clean soil matrix and acid digestion reagents used in the
extraction procedure.  A minimum of one method blank sample was analyzed for each of the 23 batches of
demonstration samples submitted for reference laboratory analysis. All method blanks provided results for
target analytes at concentrations less than 2 times the levels shown in Table 3-2.

Laboratory Control  Samples

    All LCSs met the QAPP QC requirements for all primary and secondary analytes except those
discussed below.

    The primary analytes copper and lead were observed outside the QC limits in one of the 23 batches of
samples analyzed. Reanalysis of the affected batches was not performed by the reference laboratory.
These data were qualified by the reference laboratory. Copper and lead data for all samples included in the
affected batches were rejected and not used for demonstration statistical comparisons.

    Concentrations of secondary analytes antimony, nickel, and cadmium were observed outside the QC
limits in the LCSs. Antimony LCS recoveries were continually outside the control limits, while nickel and
cadmium LCS recoveries were only occasionally outside QC limits. Antimony was a problem analyte and
appeared to be affected  by acid digestion, which can cause recoveries to fall outside control limits.
Antimony recoveries ranged from 70 to 80 percent. Since secondary analytes were not subject to the
corrective actions listed in the demonstration QAPP, no reanalysis was performed based on the LCS results
of the secondary target analytes.  These values were qualified by the reference laboratory.  All other
secondary analyte LCS  recoveries fell within the QAPP control limits.

Predigestion Matrix Spike Samples

    One predigestion matrix spike sample and duplicate were prepared by the reference laboratory for each
batch of demonstration  samples submitted for analysis. The predigestion matrix spike duplicate sample
was not required by the QAPP, but it is a routine sample prepared by the reference laboratory. This
duplicate sample can provide  data that indicates if out-of-control recoveries are due to matrix interferences
or laboratory errors.
                                               26

-------
    Predigestion spike recovery results for the primary analytes arsenic, barium, chromium, copper, lead,
and zinc were outside control limits for at least 1 of the 23 sample batches analyzed by the reference
method. These control limit problems were due to either matrix effects or initial spiking concentrations
below native analyte concentrations.

    Barium, copper, and lead predigestion matrix spike recovery results were outside control limits in
sample batches 2, 3, and 5.  In all of these cases, the unacceptable recoveries were caused by spiking
concentrations that were much lower than native concentrations of the analytes. These samples were re-
prepared, spiked with higher concentrations of analytes, reextracted, and reanalyzed.  Following this
procedure, the spike recoveries fell within control limits upon reanalysis.

    One predigestion matrix spike recovery was outside control limits for arsenic. The predigestion matrix
spike duplicate sample also was outside of control limits.  This sample exhibited an acceptable RPD for the
recovery of arsenic in the predigestion matrix spike and duplicate. A matrix interference may have been
responsible for the low recovery. This sample was not reanalyzed.

    Chromium predigestion matrix spike recoveries were  outside control limits in 7 of the 23 batches of
samples analyzed.  Five of these seven failures exhibited recoveries  ranging from 67 to 78 percent, close to
the low end of the control limits.  These recoveries were similar in the predigestion matrix spike duplicate
samples prepared and analyzed in the same batch. This indicates that these five failures were due to matrix
interferences. The predigestion matrix spike duplicate samples prepared and analyzed along with the
remaining two failures did not agree with the recoveries of the postdigestion matrix spike samples,
indicating that these two failures may be due to laboratory error, possibly inaccuracies in sample spiking.
These seven predigestion matrix spike samples were not reanalyzed.

    The zinc predigestion matrix spike recovery data were outside control limits for four batches of
samples analyzed.  In three of the spike recovery pairs, recoveries ranged from 70 to 76 percent, close to
the lower end of the control limits. The fourth recovery was much less than the lower end of the control
limits.  All  of the predigestion matrix spike duplicate samples provided recoveries that agreed with the
recoveries for the predigestion matrix spike sample recoveries indicating that the low recoveries were due to
matrix effects.  These predigestion matrix spikes and associated samples were not reanalyzed.

    The secondary analytes, cadmium, iron, and nickel, had predigestion spike recoveries outside control
limits.  Cadmium spike recoveries were outside control limits six times.  These recoveries ranged from 71
to 85 percent.  Iron spike recoveries were outside of control limits once.  Nickel spike recoveries were
outside control limits four times.  These recoveries ranged from 74 to 83 percent. Antimony spike
recoveries were always outside control limits.  No corrective  action was taken for these secondary target
analytes.

    Demonstration sample results for all target analytes that did not meet the control limits for predigestion
matrix spike recovery were qualified by the reference laboratory.

Postdigestion Matrix Spike Samples

    All postdigestion matrix spike results were within the control limit of 80 - 120 percent recovery for the
primary analytes.
    Secondary analytes, antimony, and iron were observed outside the control limits.  However, no
corrective action was taken for secondary analytes as stated in the demonstration QAPP. All postdigestion

                                                27

-------
spike recoveries for target analytes met the QA/QC requirements of the QAPP and were considered
acceptable.

Predigestion Laboratory Duplicate Samples

    Predigestion laboratory duplicate RPD results were within the control limit of 20 percent for analyte
concentrations greater than  10 times the LRL except for the following instances.  RPDs for primary
analytes barium, arsenic, lead, chromium, and copper were observed above the control limit in five
predigestion laboratory duplicate samples. These samples were reanalyzed according to the corrective
actions listed in the QAPP.  The reanalysis produced acceptable RPD results for these primary analytes.

    RPD results for the secondary analytes antimony, nickel, and cadmium were observed outside the
control limit for a number of sample batches. No corrective action was taken for secondary analytes that
exceeded the RPD control limit.

Postdigestion Laboratory Duplicate Samples

    All primary analyte postdigestion laboratory duplicate RPD results were less than the 10 percent
control limit for analyte concentrations greater than 10 times the LRL.

    The RPDs for secondary analytes antimony  and iron were observed above the 10 percent control limit
in two sample batches. No  corrective action was taken for secondary target analytes that exceeded the
RPD control limit.

Performance Evaluation  Samples

    PE samples were purchased from Environmental Resource Associates (ERA). The PE samples are
Priority PollutnT™/Contract Laboratory Program (CLP) QC standards for inorganics in soil. This type of
sample is used by the EPA to verify accuracy and laboratory performance.  Trace metal values are certified
by interlaboratory round robin analyses. ERA lists performance acceptance limits (PAL) for each analyte
that represent a 95 percent confidence interval (CI) around the certified value. PALs are generated by peer
laboratories in ERA's InterLaB™ program using the same samples that the reference laboratory analyzed
and the same analytical methods. The reported value for each analyte in the PE sample must fall within the
PAL range for the accuracy to be acceptable. Four PE samples were submitted "double blind" (the
reference laboratory was not notified that the samples were QC samples or of the certified values for each
element) to the reference laboratory for analysis by EPA SW-846 Methods 3050A/6010A. Reference
laboratory results for all target analytes are discussed later in this section.

    Four certified reference materials (CRM) purchased from Resource Technology Corporation (RTC)
also were used as PE samples to verify the accuracy and performance of the reference laboratory. These
four CRMs were actual samples from contaminated sites. They consisted of two  soils, one sludge, and one
ash CRM. Metal values in the CRMs are certified by round robin analyses of at least 20 laboratories
according to the requirements specified by the EPA Cooperative Research and Development Agreement.
The certified reference values were determined by EPA SW-846 Methods 3050A/6010A. RTC provides  a
95 percent PAL around each reference value in which measurements should fall 19 of 20 times. The
reported value from the reference laboratory for each analyte must fall within this PAL for the accuracy to
be considered acceptable. As with the four PE samples, the four CRMs were submitted "double blind" to
                                              28

-------
the reference laboratory for analysis by EPA SW-846 Methods 3050A/6010A.  The reference laboratory
results for the target analytes are discussed later in the Accuracy subsection.

Standard Reference Material Samples

    As stated in the demonstration plan (PRC  1995), PE samples also consisted of SRMs. The SRMs
consisted of solid matrices such as soil, ash, and sludge.  Certified analyte concentrations for SRMs are
determined on an analyte by analyte basis by multiple analytical methods including but not limited to ICP-
AES, flame atomic absorption spectroscopy, ICP-mass spectrometry, XRF, instrumental neutron activation
analysis, hydride generation atomic absorption spectroscopy, and polarography. These certified values
represent total analyte concentrations and complete extraction.  This is different from the PE samples,
CRM samples, and the reference methods, which use acid extraction that allows quantitation of only acid
extractable  analyte concentrations.

    The reference laboratory analyzed 14 SRMs supplied by the National Institute of Standards and
Technology (NIST), U.S. Geological Survey (USGS), National Research Council Canada, South African
Bureau of Standards, and Commission of the European Communities.  The percentage of analyses of
SRMs that were within the QAPP-defmed control limits of 80  - 120 percent recovery was calculated for
each primary and secondary analyte.

    Analyses  of SRMs were not intended to assess the accuracy of EPA SW-846 Methods 3050A/6010A
as were the ERA PE or RTC CRM  samples. Comparison of EPA SW-846 Methods 3050A/6010A acid
leach data to SRM data cannot be used to establish method validity (Kane and others 1993). This is
because SRM values are  acquired by analyzing the samples by methods other than the ICP-AES method.
In addition, these other methods use sample preparation techniques different from those for EPA SW-846
Methods 3050A/6010A.  This is one reason no PALs are published with the SRM certified values.
Therefore, the SRMs were not considered an absolute test of the reference laboratory's accuracy for EPA
SW-846 Methods 3050A/6010A.

    The SRM sample results were not  used to assess method accuracy or to validate the reference methods.
This was due to the fact that the reported analyte concentrations for SRMs represent total analyte
concentrations. The reference methods are not an analysis of total metals; rather they target the leachable
concentrations of metals. This is consistent with the NIST guidance against using SRMs to assess
performance on leaching based analytical methods (Kane and others 1993).

Data Review, Validation, and Reporting

    Demonstration data were internally reviewed and validated by the reference laboratory. Validation
involved the identification and qualification of data affected by QC procedures or samples that did not meet
the QC requirements of the QAPP. Validated sample results were reported using both hard copy and
electronic disk deliverable formats.  QC summary reports were supplied with the hard copy results.  This
qualified data was identified and discussed in the QC summary reports provided by the reference
laboratory.

    Demonstration data reported by the reference  laboratory contained three types of data qualifiers: C, Q,
and M.  Type C qualifiers included the following:

    • U - the analyte was analyzed for but not detected.
                                               29

-------
    •  B - the reported value was obtained from a reading that was less than the LRL but greater than
      or equal to the IDL.

Type Q qualifiers included the following:

    •  N - spiked sample recovery was not within control limits.

    •  * - duplicate analysis was not within control limits.

Type M qualifiers include the following:

    •  P - analysis performed by ICP-AES (Method 6010).


Quality Assessment of Reference Laboratory Data

    An assessment of the reference laboratory data was performed using the PARCC parameters discussed
in Section 2. PARCC parameters are used as indicators of data quality and were evaluated using the
review of reference laboratory data discussed above. The following sections discuss the data quality for
each PARCC parameter. This quality assessment was based on raw reference data and the raw PE sample
data.

    The quality assessment was limited to an evaluation of the primary analytes. Secondary and other
analytes reported by the reference laboratory were not required to meet the QC requirements specified in
the  QAPP. Discussion of the secondary analytes is presented in the precision, accuracy, and comparability
sections for informational purposes only.

Precision

    Precision for the reference laboratory data was assessed through an evaluation of the RPD produced
from the analysis of predigestion laboratory duplicate samples and postdigestion laboratory duplicate
samples. Predigestion laboratory duplicate samples provide an indication of the method precision, while
postdigestion laboratory duplicate samples provide an indication of instrument performance.  Figure 3-1
provides a graphical summary of the reference method precision data.

    The predigestion duplicate RPDs for the primary and secondary analytes fell within the 20 percent
control limit, specified in the QAPP, for 17 out of 23 batches  of demonstration samples. The six results
that exceeded the control limit involved only 11 of the 230 samples evaluated for predigestion duplicate
precision (Figure 3-1). This equates to 95 percent of the predigestion duplicate data meeting the QAPP
control limits.  Six of the analytes exceeding control limits had RPDs less than 30 percent.  Three of the
analytes exceeding control limits had RPDs between 30 and 40 percent. Two of the analytes exceeding
control limits had RPDs greater than 60  percent.  These data points are not shown in Figure 3-1. Those
instances where the control limits were exceeded are possibly due to nonhomogeneity of the sample or
simply to chance, as would be expected with a normal distribution of precision analyses.

    The postdigestion duplicate RPDs for the primary and secondary analytes fell within the 10 percent
control limit, specified in the QAPP, for 21 out of 23 batches  of demonstration samples. The two results
that exceeded the control limit involved only 3 of the 230 samples evaluated for postdigestion duplicate
precision in the 23 sample batches (Figure  3-1).  This equates to 99 percent of the postdigestion duplicate
data meeting the QAPP control limits. The RPDs for the three results that exceeded the control limit
ranged from 11 to 14 percent.

                                               30

-------
Relative Percent Difference (RPD)
-^ K) CO .&.
O O O O O
Predigestion Duplicate Samples


[
1
[
[
[
E
[

1
! i
: 1
]
i
j f
1

] |
] '
|
i

1
1 [
! !
]
] E
t
t
3 	 '
]
]
1 1
1
1
[
[
[
1 i
[

]
] i
i [
i 	
1
]

t
,
i
i

t
i
i [
! i

i
i
i
i
1

Antimony Arsenic Barium Chromium Cadmium Copper Iron Lead Nickel Zinc
Analyte

Relative Percent Difference (RPD)
-^ K) CO .&.
00000
Postdigestion Duplicate Samples
-
-
-
I



'
!



1


[
1


]
	 [
1 1


j 	
1 i


[
i i


]
i :



1


t
i i


i
i

Antimony Arsenic Barium Chromium Cadmium Copper Iron Lead Nickel Zinc
Analyte
    Figure 3-1.  Pre- and Postdigestion Duplicate Samples:  The top graph illustrates the
    reference laboratory's performance on analyzing predigestion duplicate samples.  Twenty
    percent RPD represents the predigestion duplicate control limits defined in the demonstration
    QAPP.  Two points were deleted from this top graph: barium at 65 percent RPD and copper at
    138 percent RPD. The  bottom graph illustrates the reference laboratory's performance on
    analyzing Postdigestion duplicate samples. Ten percent RPD represents the Postdigestion
    duplicate control limits defined in the demonstration QAPP.
Accuracy

    Accuracy for the reference laboratory data was assessed through evaluations of the PE samples
(including the CRMs), LCSs, method blank sample results, and pre- and postdigestion matrix spike
samples.  PE samples were used to assess the absolute accuracy of the reference laboratory method as a
whole, while LCSs, method blanks, and pre- and postdigestion matrix spike samples were used to assess
the accuracy of each batch of demonstration samples.

    A total of eight PE and CRM samples was analyzed by the reference laboratory.  These included four
ERA PE samples and four RTC CRM samples. One of the ERA PE samples was submitted to the
                                             31

-------
reference laboratory in duplicate, thereby producing nine results to validate accuracy. The accuracy data
for all primary and secondary analytes are presented in Table 3-3 and displayed in Figure 3-2. Accuracy
was assessed over a wide-concentration range for all 10 analytes with concentrations for most analytes
spanning one or more orders of magnitude.

    Reference  laboratory results for all target analytes in the ERA PE samples fell within the PALs. In the
case of the RTC CRM PE samples, reference laboratory results for copper in one CRM and  zinc in two
CRMs fell outside the published acceptance limits.  One of the two out-of-range zinc results  was only
slightly above the upper acceptance limit (811 versus 774 mg/kg).  The other out-of-range zinc result and
the out-of-range copper result were about three times higher than the certified value and occurred in the
same CRM.  These two high results skewed the mean percent recovery for copper and zinc shown in Table
3-3.  Figure 3-2 shows that the remaining percent recoveries for copper and zinc were all near 100 percent.

    Table 3-3 shows that a total of 83 results was obtained for the 10 target analytes. Eighty of the 83
results or 96.4  percent fell within the PALs.  Only 3 out of 83 times did the reference method results fall
outside PALs.  This occurred once for copper and twice for zinc. Based on this high percentage of
acceptable results for the ERA and CRM PE samples, the accuracy of the reference methods was
considered acceptable.

 Table 3-3.  Reference Laboratory Accuracy Data for Target Analytes
Mean Range of SD of
Percent Within Percent Percent Percent Concentration
Analyte n Acceptance Range Recovery Recovery Recovery Range (mg/kg)
Antimony
Arsenic
Barium
Cadmium
Chromium
Copper
Iron
Lead
Nickel
Zinc
6
8
9
9
9
9
7
8
9
9
100
100
100
100
100
89
100
87.5
100
78
104
106
105
84
91
123
98
86
95
120
83-125
90-160
83-139
63-93
77-101
90 - 332
79-113
35-108
79-107
79 - 309
15
22
21
10
8
79
12
22
10
72
50 - 4,955
25 - 397
19-586
1 .2 - 432
11 -187
144-4,792
6,481 - 28,664
52-5,194
13-13,279
76 - 3,021
 Notes:          n Number of samples with detectable analyte concentrations.
              SD Standard deviation.
            mg/kg Milligrams  per kilogram.

    LCS percent recoveries for all the primary analytes were acceptable in 21 of the 23 sample batches.
Lead recovery was unacceptable in one sample batch and lead results for each sample in that batch were
rejected.

    Copper recovery was unacceptable in another sample batch, and copper results for each sample in this
batch also were rejected. Percent recoveries of the remaining primary analytes in each of these two batches
were acceptable. In all, 136 of 138 LCS  results or 98.5 percent fell within the control limits.
                                               32

-------
    Method blank samples for all 23 batches of demonstration samples provided results of less than 2 times
the LRL for all primary analytes. This method blank control limit was a deviation from the QAPP, which
had originally set the control limit at no target analytes at concentrations greater than the LRL. This
control limit was widened at the request of the reference laboratory.  A number of batches were providing
method blank results for target analytes at concentrations greater than the LRL, but less than 2 times the
LRL. This alteration was allowed because even at 2 times the LRL, positive results for the method blank
samples were still significantly lower than the MDLs for each of the FPXRF analyzers. The results from
the method blank samples did not affect the accuracy of the reference data as it was to be used in the
demonstration statistical evaluation of FPXRF analyzers.

    The percent recovery for the predigestion matrix spike samples fell outside of the 80 - 120 percent
control limit specified in the QAPP in several of the 23 batches of demonstration samples.  The
predigestion matrix spike sample results indicate that the accuracy of specific target analytes in samples
from the affected batches may be suspect. These results were qualified by the reference laboratory. These
data were not excluded from use for the demonstration statistical comparison.  A discussion of the use of
this qualified data is included in the "Use of Qualified Data for Statistical Analysis" subsection.

    The RPD for the postdigestion matrix spike samples fell within the 80 - 120 percent control limit
specified in the QAPP for all 23 batches of demonstration samples.

    The QA review of the reference laboratory data indicated that the absolute accuracy of the method was
acceptable. Based on professional judgement, it was determined that the small percentage of outliers did
not justify rejection of any demonstration sample results from the reference laboratory. The accuracy
assessment also indicated that most of the batch summary data were acceptable. Two batches were
affected by LCS  outliers, and some data were qualified due to predigestion matrix spike recovery outliers.
This data was rejected or qualified.  Rejected data was not used.  Qualified data were used as discussed
below.

Representativeness

    Representativeness of the analytical data was evaluated through laboratory audits performed during the
course of sample analysis and by QC sample analyses, including method blank samples, laboratory
duplicate samples, and CRM and PE samples.  These QC samples were determined to provide acceptable
results.  From these evaluations, it was determined that representativeness of the reference data was
acceptable.

Completeness

    Results were obtained for all soil samples extracted and analyzed by EPA SW-846 Methods
3050A/6010A. Some results were rejected or qualified.  Rejected results were deemed incomplete.
Qualified results were usable for certain purposes and were deemed as complete.

    To calculate  completeness, the number of nonrejected results was determined.  This number was
divided by the total number of results expected, and then multiplied by 100 to express completeness as a
percentage. A total of 385 samples was  submitted for analysis. Six primary analytes were reported,
resulting in an expected 2,310 results. Forty of these were rejected, resulting in 2,270 complete results.
Reference laboratory completeness was determined to be 98.3 percent, which exceeded the objective for
this demonstration of 95 percent. The reference laboratory's completeness was, therefore, considered
acceptable.

                                               33

-------
    10000
                                       150
     1000 -
  D)
  _§

  C
  o
  ••=   100-
  2


  I
  o
  O
                    Antimony

   I Reference Data  CD True Value
                              IPercent Recovery
                                                        500
                                                     O)
                                                        400
                                                    O

                                                   1
                                                    CD
                                                    O
                                                    c
                                                    o
                                                   O
                                                      300
            200
            100
                                                                                            200
                                                                                            160
                                                                                            120
                                                 80
                                                                                            40
                                                                        Arsenic
                                                       •Reference Data CUTrue Value
                                        1% Recovery
CD
y
CD
Q.
    800
  $
  ~
    600
c

I 400

'E
CD

g
O
    20°
                                       150
                                       125
                                       100
                                       75
                                       50
I
CD
0.
                    Barium

     IReference Data CUTrue Value
                                 i Recovery
Concentration (mg/kg)
• -* M 00 ^ O1
O O O O O
;0 O O O O O O

~t




\




r
;ference Data C




i
c
UTru
n



If
adm
e Va

1
1
il

-
^ O) 00 -^
D O O O
O
Percent Recovery
urn
ue •% Recovery
                             200
                                                                   120
                                              Chromium

                                    IReference Data   CUTrue Value
                                    IPercent Recovery
Figure 3-2. Reference Method PE and CRM Results: These graphs illustrate the relationship between
the reference data and the true values for the  PE or CRM samples. The gray bars represent the percent
recovery for the reference data. Each set of three bars (black, white, and gray) represents a single PE or
CRM sample.  Based on this high percentage  of acceptable results for the ERA and CRM PE samples,
the accuracy of the reference laboratory method was considered acceptable.
                                                34

-------
100000-
&
^ 1 nnnn
£
c
o
~ 1 nnn
-fc=
c
I "»--(]
0 I
m


...


I HI


....
-


...

.....
Copper
•Reference Data CDTrue Value •% Recov
400
300 E1
8
0
200 <£
tr
0
£
100 £
0
ery
g Concentration (mg/kg)
ro Thousands
? - NJ U
CD O1 O1 O1 O1


1
1





r


1


§-
	
~
Iron
ce Data CUTrue Value •Percent Rec
g ^ o> oo -* -*
< o o o o M
? 00
Percent Recovery
    10000
  o
  •-§
  i=
  CD
  O
  c
  o
  O
     1000
100
       10
1
           I
                                       125
                                       100
                                 75
                                 50
                                       25
CD
CL
                      Lead

           IReference Data  CUTrue Value
           IPercent Recovery
                                                       100000
                                                                                          120
                                                                   Nickel

                                                    •Reference Data CDTrue Value
                                                                                    40
                                                                                Recovery
                           100000
                            10000
                        •     1 ooo
                  "c
                  CD

                  I
                  o
                              100
                               10
                                                               400
                                                               300  £•
                                                               200
                                                                100
                                                                    O
                                                                    O
                                                                    CD
                                                                    "c
                                                                    CD
                                                                    O
                                              Zinc
                             •Reference Data CUTrue Value
                                                  1% Recovery
Figure 3-2 (Continued). Reference Method PE and CRM Results:  These graphs illustrate the
relationship between the reference data and the true values for the PE or CRM samples.  The gray bars
represent the percent recovery for the reference data.  Each set of three bars (black, white, and gray)
represents a single PE or CRM sample.  Based on this high percentage of acceptable results for the ERA
and CRM PE samples, the accuracy of the reference laboratory method was considered acceptable.
                                               35

-------
Comparability

    Comparability of the reference data was controlled by following laboratory SOPs written for the
performance of sample analysis using EPA SW-846 Methods 3050A/6010A.  QC criteria defined in the
SW-846 methods and the demonstration plan (PRC 1995) were followed to ensure that reference data
would provide comparable results to any laboratory reporting results for the same samples.

    Reference results indicated that EPA SW-846 Methods 3050A/6010A did not provide comparable
results for some analytes in the SRM samples.  SRM performance data for target analytes is summarized in
Table 3-4 and displayed in Figure 3-3. As with the PEs, the analyte concentrations spanned up to 3 orders
of magnitude in the SRMs.  The percentage of acceptable (80 - 120 percent recovery) SRM results  and
mean percent recovery was less than 50 percent for the analytes antimony, barium, chromium, iron, and
nickel. The low recoveries for these five analytes reflect the lesser tendency for them to be acid-extracted
(Kane and others 1993).

    Under contract to the EPA, multiple laboratories analyzed NIST SRMs 2709, 2710, and 2711 by EPA
SW-846 Methods 3050A/6010A. A range, median value, and percent leach recovery based on the median
value for each detectable element were then published as  an addendum to the SRM certificates.  These
median values are not certified but provide a baseline for comparison to other laboratories analyzing these
SRMs by EPA SW-846 Methods 3050A/6010A.  Table 3-5 presents the published percent leach recovery
for the 10 primary and secondary analytes and the reference laboratory's results for these three NIST
SRMs. Table 3-5 shows that the results produced by the  reference laboratory were consistent with the
published results indicating good comparability to other laboratories using the same analytical methods on
the same samples.

 Table 3-4.  SRM Performance Data for Target Analytes
Percent Within Mean Range of SD of
Acceptance Percent Percent Percent Concentration
Analyte n Range Recovery Recovery Recovery Range (mg/kg)
Antimony
Arsenic
Barium
Cadmium
Chromium
Copper
Iron
Lead
Nickel
Zinc
5
11
8
10
10
17
7
17
16
16
0
72
12
50
0
88
14
82
19
75
22
84
41
80
45
82
62
83
67
81
15-37
67- 106
21 -89
43-95
14-67
33-94
23-84
37-99
25-91
32-93
9
10
21
15
16
17
25
17
17
14
3.8- 171
18-626
414- 1,300
2.4 - 72
36 - 509
35 - 2,950
28,900 - 94,000
19-5,532
14-299
81 - 6,952
 Notes:            n Number of SRM samples with detectable analyte concentrations.
                 SD Standard deviation.
              mg/kg Milligrams per kilogram.
                                              36

-------
 Table 3-5. Leach Percent Recoveries for Select NIST SRMs
 Analyte
                     NIST SRM 2709
              Reference
Published    Laboratory
 Result3        Result
                                NIST SRM 2710
              Reference
Published     Laboratory
 Result3        Result
                                NIST SRM 2711
              Reference
Published     Laboratory
 Result3        Result
Antimony
Arsenic
Barium
Cadmium
Chromium
Copper
Iron
Lead
Nickel
Zinc
-
-
41
-
61
92
86
69
89
94
-
106
37
-
-
85
84
87
76
78
21
94
51
92
49
92
80
92
71
85
-
87
45
84
-
92
78
96
69
88
-
86
28
96
43
88
76
95
78
89
20
91
25
87
49
90
66
90
70
85
 Notes:            Published results found in an addendum to SRM certificates for NIST SRMs 2709, 2710, and
                  2711.
            NIST  National Institute of Standards and Technology.
            SRM  Standard reference materials.
               -  Analyte not present above the method LRL.


    The inability of EPA SW-846 Methods 3050A/6010A to achieve the predetermined 80 - 120 percent
recovery requirement indicated that the methods used to determine the certified values for the SRM samples
were not comparable to EPA SW-846 Methods 3050A/6010A.  Differences in the sample extraction
methods and the use of different analytical instruments and techniques for each method were the major
factors of this noncomparability. Because of these differences, it was not surprising that the mean percent
recovery was less than 100 percent for the target analytes. The lack of comparability of EPA SW-846
Methods 3050A/6010A to the total metals content in the SRMs did not affect the quality of the data
generated by the reference laboratory.

    The assessment of comparability for the reference data revealed that it should be comparable to other
laboratories performing analysis of the same samples using the same extraction and analytical methods, but
it may not be comparable to laboratories performing analysis of the same samples using different extraction
and analytical methods or by methods producing total analyte concentration data.

Use  of Qualified Data  for Statistical Analysis

    As noted above, the reference laboratory results were reported and validated, qualified, or rejected by
approved QC procedures.  Data were qualified for predigestion matrix spike recovery and pre- and
postdigestion laboratory duplicate RPD control limit outliers.  None of the problems were considered
sufficiently serious to preclude the use of coded data.  Appropriate corrective action identified in the
demonstration plan (PRC 1995) was instituted. The result of the corrective action indicated that the poor
percent recovery and RPD results were due to matrix effects.  Since eliminating the matrix effects would
require additional analysis using a different determination method such as atomic absorption spectrometry,
or the method of standard addition, the matrix effects were noted and were not corrected.
                                               37

-------
    PARCC parameters for the reference laboratory data were determined to be acceptable. It was
expected that any laboratory performing analysis of these samples using EPA SW-846 Methods
3050A/6010A would experience comparable matrix effects. A primary objective of this demonstration was
to compare sample results from the FPXRF analyzers to EPA SW-846 Methods 3050A/6010A, the most
widely used approved methods for determining metal concentrations in soil samples. The comparison of
FPXRF and the reference methods had to take into account certain limitations of both methods, including
matrix effects.  For these reasons, qualified reference data were used for statistical analysis.

    The QC review and QA audit of the reference data indicated more than 98 percent of the data either
met the demonstration QAPP objectives or was QC coded for reasons not  limiting its use in the data
evaluation. Less than 2 percent of the data were rejected based on QAPP  criteria. Rejected data were not
used for statistical analysis.  The reference data were considered as good as or better than other laboratory
analyses of samples performed using the same extraction and analytical methods. The reference data met
the definitive data quality criteria and was of sufficient quality to support  regulatory activities.  The
reference data were found to be acceptable for comparative purposes with the FPXRF data.
Concentration (mg/kg)
" o a s a s
yin
n


J
n


.




D -^ K) CO J
O O O C
Percent Recovery
Antimony
ference Data d True Value • Percent Recovery

Concentration (mg/kg)
• — ^ o o
-^ o o o
^j O O O O


-
MM f
	
Barium
eference Data CDTrue Value •% Recovery
100
80 >,
0
60 8
0
ce
40 |
0
20 °-
0



Concentration (mg/kg)

1



i
J
Bference Data
It,


.


r


\


Arsenic
D True Value •Percent Rec
120
100 £•
0
>
o
o
0
so o:
-I-J
C
60 |
40
wery

on
Concentration (mg/kg)
K) ^. O) C
D O O O C










f


g-






f t

"••*
Cadmium
• Reference Data DTrue Value •% Recovery
K) ^. O) 00 -^
O O O O O
Percent Recovery
  Figure 3-3. Reference Method SRM Results: These graphs illustrate the relationship between the
  reference data and the true values for the SRM samples. The gray bars represent the percent
  recovery for the reference data.  Each set of three bars (black, white, and gray) represents a single
  SRM sample.
                                               38

-------
    600
    400
    200
o
c
o
O
                                          80
                                        60 £•
                                           0
                                           >
                                           o
                                           o
                                           0
                                        40 o:
                                           -I-J


                                           §

                                        20 «
                    Chromium


       I Reference Data Dime Value
                                1% Recovery
                                                     10000
                                                 If

                                                  E  1000
                                                   c
                                                   o
                                                   0
                                                   o
                                                   c
                                                   o
                                                   O
                                                       100
                                                        10
                                                                                            100
                                                                      Copper
                                                       I Reference Data  dime Value
                                                                                 • Percent Recovery
       110
                                         100
II
  c
  o
  O
                                       20
                                             0


                                             8
                                             0
                                             OL

                                             'c
                                             0
                                             S
                                             0
                                             Q_
                                                     10000
                                                 |  1000


                                                  c
                                                  o
                                                  8

                                                  o
                                                 O
                                                      100
                                                        10
                                                                                            120
                        Iron


        I Reference Data dime Value
                                 1% Recovery
                                                                       Lead


                                                       I Reference Data dime Value
                                                                                 l%Recovery
    400
| 300





E 200
c


c  100
o
O
100




80




60




40




20




0
                                                     10000
                                                   c
                                                   o
                                                   •
8

o
O
                                                      1000
                                                       100
                                                        10
                                                                                            100
                                                                                           80
                                                                                           60
                                                                                           40
                                                                                           20
                                                                                              0


                                                                                              8
                                                                                              0
                                                                                              OL

                                                                                              'c
                                                                                              0
                                                                                              Q

                                                                                              0
                                                                                              Q.
                     Nickel



      I Reference Data dime Value
                               1% Recovery
                                                                       Zinc



                                                       I Reference Data dime Value
                                                                                 1% Recovery
Figure 3-3 (Continued).  Reference Method SRM Results:  These graphs illustrate the relationship

between the reference data and the true values for the SRM samples.  The gray bars represent the

percent recovery for the reference data.  Each set of three bars (black, white, and gray) represents a

single SRM sample.
                                                39

-------
                                          Section 4
                                X-MET 920-MP Analyzer
    This section provides information on Metorex's X-MET 920-MP Analyzer, including the theory of
FPXRF, operational characteristics, performance factors, a data quality assessment, and a comparison of
its results with those of the reference laboratory.

Theory of FPXRF Analysis

    FPXRF analyzers operate on the principle of energy dispersive XRF spectrometry. This is a
nondestructive qualitative and quantitative analytical technique that can be used to determine the metals
composition in a test sample. By exposing a sample to an X-ray source having an excitation energy close
to, but greater than, the binding energy of the inner shell electrons of the target element, electrons are
displaced. The electron vacancies that result are filled by electrons cascading in from outer shells.
Electrons in these outer shells have higher potential energy than inner shell electrons, and to fill the
vacancies, the outer shell electrons give off energy as they cascade into the inner shell (Figure 4-1). This
release of energy results in an emission of X-rays that is characteristic to each element. This remission of
X-rays is termed XRF.

    Because each element has a unique electron shell configuration, each will emit unique X-rays at fixed
wavelengths called "characteristic" X-rays. The energy of the X-ray is measured in electron volts (eV). By
measuring the peak energies of X-rays emitted by a sample, it is possible to identify and quantify the
elemental composition of a sample.  A qualitative analysis of the sample can be made by identifying the
characteristic X-rays produced by the sample. The intensity of each characteristic X-ray emitted is
proportional to the concentration of a given element and can be used to quantitate each analyte.

    Three electron shells are generally involved in the emission of characteristic X-rays during FPXRF
analysis: the K, L, and M shells. A typical emission pattern, also called an emission spectrum, for a given
element has multiple peaks generated from the emission X-rays by the K, L, or M shell electrons. The most
commonly measured X-ray emissions are from the K and L shells; only elements with an atomic number of
58 (cerium) or greater have measurable M shell emissions.

    Each characteristic X-ray peak or line  is defined with the letter K, L, or M, which signifies which shell
had the  original vacancy and by a subscript alpha (a) or beta (B), which indicates the next outermost shell
from which electrons fell to fill the vacancy and produce the X-ray.  For example, a Ka-line is produced by
a vacancy in the K shell filled by an L shell electron, whereas a KB-line is produced by a vacancy in the K
shell filled by an M shell electron.  The Ka transition is between 7 and 10 times more probable than the KB
transition. The Ka-line is approximately 10 times more intense than the KB-line for a given element, making
the Ka-line analysis the preferred choice for quantitation purposes.

                                               40

-------
                               Bccitation X-ray from the
                                  FPXRF Source
                                         N,
                                           N.
                                             X
      An excited electron is displaced, creating an
              electron vacancy.
    An outer electron shell electron cascades to the inner electron shell to
    fill the vacancy. As this electron cascades, it releases energy in the
                    form of an X-ray.
                                    Characteristic X-ray

  Figure 4-1. Principle of Source Excited X-ray Fluorescence: This figure illustrates the dynamics
  of source excited X-ray fluorescence.

    For a given element, the X-rays emitted from L transitions are always less energetic than those emitted
from K transitions.  Unlike the K-lines, the L-lines (La and LB) for an analyte are of nearly equal intensity.
The choice of which one to use for analysis depends on the presence of interfering lines from other analytes.

    An X-ray source can excite characteristic X-rays from an analyte only if its energy is greater than the
electron binding energies for the target analyte.  The electron binding energy, also  known as the absorption
edge energy,  represents the amount of energy an electron has to absorb before it is displaced. The
absorption edge energy is  somewhat greater than the corresponding line energy. Actually, the K-absorption
edge energy is approximately the sum of the K-, L-, and M-line  energies of the particular element, and the
L- absorption edge energy is approximately the sum of the L- and M-line energies. FPXRF analytical
methods are more sensitive to analytes with absorption edge energies close to, but less than, the excitation
energy of the source.  For example, when using a Cd109 source, which has an excitation energy of 22.1
kiloelectron volts (keV), an FPXRF analyzer would be more sensitive to zirconium, which has a K-line
absorption edge energy of 15.7 keV, than to chromium, which has a K-line absorption edge energy of 5.41
keV.

Background

    Metorex  is an international supplier of advanced equipment for metal detection, materials testing, and
chemical analysis.  It offers a wide range of products from field portable and laboratory-grade metals and
alloy analyzers to in-line (process) analyzers. It has more than 20 years of experience in developing X-ray
analyzer technologies.

    Metorex  developed the X-MET 920 line as a modular system, which can be configured with the
hardware and software needed to address a specific analysis problem.  With the X-MET 920 line, Metorex
                                                 41

-------
offers several analysis probes. The probe contains the excitation sources and either a silicon (drifted with
lithium) Si(Li) detector or a gas-filled proportional counter detector.

    Metorex developed the X-MET 920 line of instruments to perform elemental analysis in the petroleum
and petrochemical industry, the mining and minerals industry, and the environmental field.  The X-MET
920-MP is a field portable technology that can be operated in the in situ or intrusive mode. The X-MET
920-MP is Metorex's miniportable version on the X-MET 920 line. The analyzer uses energy dispersive
XRF spectroscopy to  determine elemental composition of soils and other solid waste materials as well as
liquids and slurries. The X-MET 920-MP can identify and quantify 32 elements simultaneously, and 70
elements can be identified and quantified when all available sources are used. Metorex offers four
excitation sources, iron-55  (Fe55), cadmium-109 (Cd109), americium-241 (Am241), and curium-244 (Cm 244),
and two detectors, [Si(Li) and gas-filled proportional counter] in its probes.  For this demonstration, the
analyzer used the surface analysis probe system (SAPS) equipped with a Cd109 source and a gas-filled
proportional detector. The SAPS is designed to only house one excitation source.  Metorex also offers a
dual source version of the SAPS probe called the double source probe system (DOPS). The choice of the
source is determined by the user depending on the target analytes.

    For in situ analyses, the probe is pointed downward against the soil  surface to allow the source-
detector window to come into contact with the soil surface.  For intrusive analyses, the probe is pointed
upward and a protective sample cover is attached over the probe window.  The soil samples are  placed in
sample cups inside the protective sample cover for analysis. In either mode, sample measurement is
initiated by depressing a trigger built into the handle of the probe. This exposes the sample to primary
radiation from the excitation source.  Fluorescent and backscattered radiation reenters the probe through
the probe window and is counted by the detector, which then transmits the electronic signal to the MCA for
processing. The X-MET 920-MP is operated and calibrated using the X-MET software, which uses an
empirical calibration.

Operational Characteristics

    This section discusses equipment and accessories, operation of the analyzer in the field, background of
the operator, training, reliability of the analyzer, health and safety concerns, and representative operating
costs.

Equipment and Accessories

    The X-MET 920-MP comes with all the equipment necessary for both in situ and intrusive operation.
Metal, foam-padded carrying cases are provided for storage of the XPCS unit and the analysis probe.
Specifications for the X-MET 920-MP are provided in Table 4-1.

    Three main components comprise the analytical system: a laptop computer, the electronics unit or
portable XPCS, and the probe.  The XPCS contains a 2,048-channel MCA that is used to collect, analyze,
and display the data.  It is housed in a rugged, weatherproof, self-contained unit. The portable XPCS is
about the size of a laptop computer and weighs approximately 5 pounds. It can be powered from 110 or
220V AC electricity or from 10 internal C-sized rechargeable nickel-cadmium batteries. With a full
charge, the batteries last 7 to 8 hours before a recharge is needed. The XPCS comes standard with 10
rechargeable batteries and a charger unit and is equipped with a cable that interfaces to an RS-232 port of
a laptop computer.  It also contains a port for the cable that plugs into the  probe.
                                               42

-------
 Table 4-1. Analyzer Instrument Specifications
Characteristic Specification
Resolution
Source
Detector
Probe Size
Probe Weight
XPCS Size
XPCS Weight
XPCS Data Acquisition
XPCS Communication
Computer Requirements
Power Source
Operational Checks
Intrusive Operation
Contact: Jim Pasmore
1900 NE Division Street
Bend, OR 97701
1-800-229-9209
(503) 385-6750 (FAX)
760 eV (Manganese-Ka)
10 mCi Cd109 (Fe55, Am 241, and Cm244 also available)
Gas-filled Proportional (Si(Li) also available)
20 cm x 25 cm x 5 cm
2.3 kilograms (kg)
28 cm x 23 cm x 4.5 cm
2.3 kg (with batteries)
2,048-channel MCA
Via RS-232 port at baud rate of 38.4 kilobytes
IBM compatible PC with MS DOS, 486 DX CPU, 4 megabyte
random access memory, 170 MB hard disk, VGA graphics, RS-
232 port and RS-232 cable, and parallel printer
120V or 220V alternating current or batteries
Gain check and site-specific check sample once per hour
Attachment of protective sample cover

    The X-MET 920-MP uses the laptop computer's RS-232 port to interface directly with the portable
XPCS. The requirements of the laptop computer are also listed in Table 4-1.  The laptop computer can be
supplied by Metorex or by the user.  If Metorex supplies the laptop computer, the X-MET software will
already be installed. If the user supplies the laptop computer, Metorex will provide instructions to install
the software.  During this demonstration, Metorex supplied a laptop computer with the X-MET software
already installed. PRC provided two more laptop computers, one owned and one rented.  Metorex helped
the operator install the X-MET software on both. All three computers were used during the demonstration.
As the data is collected and processed by the portable XPCS, the spectra and results files  are  saved on the
hard drive of the laptop computer. The data can also be sent to a printer for a hard copy printout.

    The probe used with the X-MET 920-MP Analyzer during this demonstration was the SAPS probe. It
weighed approximately 3 pounds and used a 10 millicurie (mCi) Cd109 radioisotope source for excitation of
the target analytes.  The Cd109 source was installed in March 1995, 1 month before the demonstration.  The
radioisotope source was in the form of an 8-mm-diameter by 5-mm-thick capsule. The probe's beryllium
sample window is environmentally sealed by a 25-mm-diameter window of clear Kapton™ film.

    The SAPS probe contains a gas-filled proportional detector.  This detector is a chamber filled with a
noble gas and fitted with a central wire electrode biased to several hundred volts. An X-ray photon entering
the chamber ionizes the gas atoms. The electric charge produced is collected and provides an electric pulse
                                               43

-------
whose magnitude (amplitude) is directly proportional to the energy of the X-ray photon absorbed by the gas
in the detector.  This detector is rugged and lightweight, it operates at ambient temperatures from -13 to
140 °F.  The detector achieves a manganese Ka X-ray resolution of 760 eV. This detector has a lower
resolution than the Si(Li) detector (170 eV); however, it has a higher efficiency because it has a larger
collection area than the Si(Li) detector. The sensitivity of the detector is the sum of the resolution and
efficiency.

    The SAPS can be used as a surface probe to perform in situ analyses by pointing downward at the soil
surface, or with the attachment of a protective sample cover, the probe can analyze soil samples intrusively
from a sample cup. The protective sample cover attaches to the probe over the sample window. Sample
measurements are initiated by depressing a trigger built into the handle of the probe. The probe is
connected to the XPCS with a coiled, flexible cord that is about 6 feet long.

    Metorex used pure element samples for the calibrations conducted at each site. To perform either an
empirical or a fundamental parameters (FP) calibration, a pure element standard for each target analyte is
required. These pure element standards are in the shape of a coin that is about the size of a $0.50 piece.
They are large enough to completely cover the probe's Kapton™ window.  Metorex also brought known
calibration standards to the test sites. These standards included several NIST SRMs and  soil samples that
were analyzed by Metorex. Metorex also brought the predemonstration split samples to each site. Other
equipment and supplies that were helpful when using the X-MET 920-MP, and are not supplied by the
developer, include protective  gloves, paper towels, a marking pen, a portable generator, and an extension
cord.  The generator and extension cord were used because of battery problems associated with the laptop
computers.

Operation of the Analyzer

    The operation of the X-MET 920-MP can be described in three  steps:  (1) performing the calibration,
(2) taking measurements, and (3) managing the data.  The X-MET 920-MP can be calibrated empirically
or through the use of FPs. In this demonstration, the X-MET 920-MP was used with the  SAPS probe
containing a gas-filled proportional detector. When using this detector, only the empirical calibration can
be performed. The FP calibration relies on the ability of the detector to fully resolve the peaks  from
neighboring elements. The gas-filled proportional detector cannot adequately resolve these peaks;
therefore, the empirical calibration must be used. A separate empirical calibration was performed at each
site.

    The empirical calibration is performed using the "Calibration" menu of the X-MET software.  Metorex
recommends that 10 - 20 site-specific calibration standards (SSCS) be used to perform the empirical
calibration. The SSCSs have known analyte concentrations usually determined by analytical methods such
as inductively coupled plasma or atomic absorption spectroscopy. If SSCSs are not available,  site-typical
calibration samples that closely approximate the site's soil matrix with respect to particle size distribution,
mineralogy, and contaminant analytes can be used.

    The first step in the empirical calibration is to analyze the pure element samples for the target elements
plus any elements that may cause interferences at a measurement time of 200 live-time  seconds. This
enables the X-MET software  to establish channel limits or windows for each element for spectral
deconvolution.  The X-MET  software automatically chooses the Ka  or La peak to set the window for the
analytes. However, it is possible for the operator to manually choose a different window or peak.  For
example, in this demonstration, the Lp peak for lead was chosen because it is almost as intense  as the La

                                               44

-------
lead peak but is not influenced by the interference from arsenic. The copper and zinc windows also were
manually edited to reduce the amount of spectral overlap.

    The next step in the empirical calibration is to enter the analyte concentrations for each SSCS into the
X-MET software. After this is completed, each SSCS was analyzed using a count time of 200 live-time
seconds. At the ASARCO site, 20 SSCSs were used and 14 SSCSs were used at the RV Hopkins site for
the empirical calibration. The SSCSs samples were collected during the predemonstration and sent to
Metorex. The analyte concentrations in these samples had been determined by EPA SW-846 Methods
3050A/6010A.

    The final step in the empirical calibration is to develop the calibration curves for each analyte of
interest. The calibration equation is developed using a multivariable, linear least squares fit regression
analysis. After the regression terms to be  used in the equation is defined, the software is designed to
provide a mathematical equation to calculate the analyte concentration in an unknown sample.  The
equation is then stored in memory and used to quantitate an unknown sample when needed.

    Once the calibration is complete, it is  permanently stored as a method in the software. To analyze
samples, the operator opens the stored method with the correct calibration and sets the measurement time.
For this demonstration, it was suggested by Metorex to use shorter count times when analyzing the in situ
samples and longer count times when analyzing the intrusive samples. This reflected the developer's
opinion that increased sample preparation should be mirrored by increased analyzer accuracy or precision.
The count times used at the ASARCO site were 30 seconds for in situ samples, 60 seconds for intrusive-
unprepared samples, and 120 seconds for  intrusive-prepared samples.  The count times used at the RV
Hopkins site were 60 seconds for the in situ samples, 120 seconds for the intrusive-unprepared samples,
and 180 seconds  for the intrusive-prepared samples. The count times were longer at the RV Hopkins site
because chromium was an analyte of interest at RV Hopkins and Metorex was concerned that longer count
times would be necessary to quantitate for chromium. The analysis is started by depressing the trigger in
the handle of the  SAPS probe. When the  analysis is complete the X-MET 920-MP beeps and the analytical
results are displayed on the screen.

    The third step in the operation of the X-MET 920-MP is managing the data.  All the analytical  data
files are automatically saved under a file name specified in the method. The raw spectrum of each analysis
can also be saved individually.  The operator noted it would be helpful if the software had a function to
automatically save the raw  spectra as it did the data files so that it would be unnecessary to manually save
each one. All results files and spectra files were backed up on a floppy disk and were printed from the
DOS editor.

Background of the Technology Operators

    Metorex made no recommendation as to the qualifications and background necessary to operate its
instrument, leaving the decision to the discretion of the client.  Two PRC employees operated the X-MET
920-MP during the demonstration.  Both operators have received all required Occupational Safety and
Health Administration training to work on sites containing hazardous wastes.

    The primary  operator during this demonstration was employed as a geologist by PRC for 4 years. He
earned a bachelor's degree in geology in 1982 and a master's degree in geology in 1987. While earning his
masters, he worked for an environmental testing laboratory. He has extensive experience in field analysis;
he worked on the Field Investigative Team contract for 3 years as a geologist and field analyst and has

                                               45

-------
managed several comprehensive site investigations.  At the conclusion of the field work at the RV Hopkins
site, some intrusive samples from both sites remained for analysis, but the original operator was not able to
finish the samples.  At this time, the original operator trained another person to finish the analyses.

    Approximately 80 of the intrusive-prepared samples from the ASARCO site and 50 of the intrusive-
prepared samples from the RV Hopkins site were analyzed by the second operator.  The replacement
operator had been employed by PRC for 6 months as an environmental scientist prior to the demonstration.
Prior to coming to PRC, he worked for 2.5 years in hazardous waste management for a Kansas
municipality. He served as a nuclear, biological, and chemical (NBC) defense specialist in the United
States Marine Corps for 6 years, specializing in monitoring and surveying of NBC attacks using field
screening technologies,  such as mobile mass spectrometers. He earned a bachelor's degree in 1992.

Training

    The original operator received 2 days of training by the developer at the start of the demonstration.
Approximately one-half day of training was dedicated to the theoretical background of XRF technology and
the other one-and-a-half days were spent on specific operation and hands-on training for the X-MET 920-
MP. The hands-on training covered performing the test measurements using the pure element samples,
empirical calibrations and FP calibrations, the analysis of various standards and soil samples, and the steps
for saving spectra and quantitative results.  The developer indicated that the standard training course is 3
days and stated that the  PRC operator received more one-on-one instruction in 2 days than typical students
in a standard 3-day course. Metorex tailors its training course to match the  level of operator experience
and the primary intended use of its equipment.

    Two Metorex representatives stayed at the ASARCO site throughout the first day of in situ analysis
and assisted the PRC operator that evening with downloading the spectra and results files to floppy disk
and printing out the data.  They departed the next morning. Another Metorex representative also came to
the RV Hopkins site for 2 days prior to the initiation of the field work to assist the operator with the
empirical calibration of the X-MET 920-MP.

Reliability

    More than 1,800 individual measurements were collected with the X-MET 920-MP Analyzer using the
SAPS probe. This included the measurement of soil samples using the four sample preparation techniques,
10 replicate measurements on 48 samples for a precision assessment, and the measurement of QC samples
such as blanks, check samples, PE samples, and SRMs.  During the demonstration, there were frequent
light to moderate rains while the analyzers were performing the in situ measurements. The temperatures
fluctuated between 5 and 16 °C at the ASARCO site and 6 and 22 °C at the RV Hopkins site.  Despite the
less than ideal weather conditions, there were no mechanical problems experienced with the X-MET
analyzer during the demonstration.  The only maintenance necessary was to  wipe the detector window with
a paper towel between readings during the in situ analyses and occasionally to wipe dust off the window
during the intrusive analyses.  The operator did note that most commercially available and economically
priced laptop computers are not weatherproof or ruggedized.  The computer was covered with a plastic
sheet while conducting the in situ measurements. Other than the inconvenience, this procedure proved
effective and the computers operated without problems.

    The primary problems encountered with the X-MET 920-MP were with battery lifetimes and operator
familiarity with the software.  The operator found that the batteries in the XPCS unit provided 7 to 8 hours

                                               46

-------
of power which was adequate for in situ analyses.  The limiting factor, however, was the battery life of the
laptop computers. The operator was only able to get about 1.5 to 2 hours of power from the computer
batteries, despite using three different computers from two manufacturers and obtaining three new batteries
and a charger for one computer.  The need to constantly change computers resulted in significant
downtime.  To keep pace with the demonstration schedule, a 35-pound portable generator was used at both
sites to supply power to the laptop computer. The use of the generator decreased throughput and ease of
use of the X-MET 920-MP.  The limited PC battery lifetime  and nonweatherproof design of the laptop
computers diminished the effectiveness of the X-MET 920-MP as an in situ FPXRF analyzer.

    The operator noted that several days were required to become familiar with the various menu options.
The operator did feel that some aspects of the software were not self evident and were somewhat awkward.

    A series of minor software problems was encountered by the operator. For example, every time
batteries were switched, the operator was locked out of the software. After having this problem two or
three times, the solution was to turn off the entire system and reinitialize the XPCS and the  computer when
changing the battery.

    It was also discovered that attempting to use the same data file after reinitialization caused the file to be
corrupted. The operator attempted to solve the problem with the assistance of Metorex but without
success.  A large portion of the data collected on the third day of analysis was never retrieved. The
operator failed to back up this data from the hard drive.  The rental laptop computer was returned before
another attempt was made to retrieve the data. Ultimately, 78 data points from the in s/YM-unprepared
sample measurements were lost. The operator learned that a new file had to be created for the data each
time a lockout occurred to avoid corrupting the original data file.

    Viewing and saving spectra also proved to be a problem at the ASARCO site. The operator was
locked out of the software several times while attempting to save a spectrum.  He had to reboot the
computer and start a new data file when this occurred. This problem often occurred when the operator
tried to type while the software was changing prompt messages.

    On the final day of field analysis at the ASARCO site while analyzing the intrusive-prepared samples,
the operator encountered an error message while saving data. The operator was concerned  about losing
data, so he copied all the spectra onto a floppy disk. He failed to back up the  data files on floppy disk and
did not print out a hard copy of this data.  This rental laptop computer was sent back to the developer
before the data was retrieved from the hard drive. This caused the loss of 13 data points from the intrusive-
prepared sample measurements at the ASARCO site.

Health and Safety

    Exposure to radiation  from the excitation source was the largest health and safety consideration while
using the analyzer.  Radiation was monitored with a gamma ray detection survey meter.  Background
radiation at the two sites was between 0.006 and 0.012 millirems per hour (mrem/hr). Radiation was
monitored in the in situ and intrusive modes while the probe's source was exposed (during a measurement),
obtaining a worst-case scenario.  The radiation was measured within 5 cm of the probe face while
analyzing a sample.  Radiation exposure also was monitored at a point on the probe where the operator's
hand was located during analysis to provide a realistic value  of operator exposure. The permissible
occupational exposure in Kansas is 5,000 millirems per year, which equates to approximately 2 to 3
mrem/hr assuming constant exposure for an entire work year.

                                               47

-------
    While taking in situ measurements (probe pointing down), a maximum radiation value of 0.60 mrem/hr
at the probe face was obtained with the Cd109 source exposed.  The radiation values dropped off to 0.03 to
0.04 mrem/hr at the probe handle when the Cd109 source was exposed. Metorex designed the SAPS probe
primarily for in situ measurements.

    Two different protective covers were used on the SAPS probe with the probe pointing upward to
conduct intrusive measurements. The SAPS was not designed to perform analysis of samples in a cup;
however, for this demonstration the developer was interested in evaluating the SAPS in the intrusive mode.
The first protective cover simply sat over the probe window but did not attach to the probe. With the
protective cover in place and a sample cup over the window, radiation values of 2.0 to 3.0 mrem/hr were
obtained at the sides of the cover with the source exposed. The radiation values dropped off to 0.10
mrem/hr at 1 foot from the probe and at the handle on the probe. Metorex sent the operator a second
protective cover that was designed to attach to the  SAPS probe. However, this protective cover was open
on one side so that a sample cup could be placed over the probe window. With the source exposed,
radiation values of 60 - 80 mrem/hr were obtained 1 inch from the open side of the protective cover. One
foot away from the open side of the protective cover, radiation values of 2.0 to 4.0 mrem/hr were recorded.
At the probe handle, 0.20 - 0.25 mrem/hr was measured with this protective cover. Based on these
radiation values, the operator chose to use the first protective cover. At the RV Hopkins site, the operator
placed a cardboard box lined with lead foil around the probe to offer additional protection.  Radiation
values of 0.016 to 0.018 mrem/hr were recorded behind the lead foil with the Cd109 exposed.

Cost

    At the time of the demonstration, the X-MET 920-MP and SAPS probe with the Cd109  radioisotope
source cost $36,325.  This includes the X-MET software, eight pure element standards, and 3 days of
training for two people at Metorex. Travel and accommodation costs for the training are not included.  The
purchase price is $3,000 less if the user supplies the laptop computer.  The cost of the miniportable XPCS
unit is $14,745. This price does not include the computer, probe, or source. Periodic maintenance includes
replacement and disposal of the 10 mCi Cd109 source every 2 to 3 years at a cost of $3,725  with an
additional $500 source disposal fee.

    The X-MET 920-MP can be leased from Metorex. There is a 1-month minimum rental. The cost is  10
percent of the purchase price per month, and all shipping costs. Metorex mandates that the users be trained
on the X-MET 920-MP.  Users have a choice of training options.  A 3-day class is offered at Metorex's
facility for $685 per person plus travel and lodging expenses. On-site training classes are also available.

    The primary cost benefit of field analysis is the quick access to analytical data.  This allows the
process dependent on the analytical results to move efficiently onto the next stage.  Costs associated with
field analysis are very dependent on the scope of the project. Since most of the mobilization costs are fixed,
analyzing a large number of samples lowers the per sample cost. This is a key advantage that field analysis
has over a conventional laboratory. Furthermore, more samples are usually taken for field  analysis since
questions raised in the preliminary  findings may be resolved completely without the need to return for
another sample collection event.
                                               48

-------
    A representative list of costs associated with the X-MET 920-MP is presented in Table 4-2. Also
included in this table is the measured throughput and the per sample charge of the reference laboratory.
Given the special requirements of this demonstration, it was not considered reasonable to report a per
sample cost for the field analysis. However, some estimates can be derived from the data provided in this
table.

          Table 4-2.  Instrument and Field Operation Costs
Item Amount
X-MET 920-MP
Replacement Source
Operator Training (Vendor Provided)
Radiation Safety License (State of Kansas)
$ 36,325
3,650
3,725
685
500
Purchase Price
Per Month Lease
For Cd109
—
—
Field Operation Costs
Supplies and Consumables (Sample cups,
window film, sieves, standards)
Field Chemist (Labor Charge)
Per diem
Travel
Sample Throughput
Cost of Reference Laboratory Analysis
300 - 500
100-150
80-120
200 - 500
8-14
150
(Varies, depending on
sample load)
Per day
Per day
Per traveler
Samples per hour
Per sample
Performance Factors

    The following paragraphs describe performance factors, including detection limits, sample throughput,
and drift.

Detection Limits

    MDLs, using SW-846 protocols, were determined by collecting 10 replicate measurements on site-
specific soil samples with metals concentrations 2 to 5 times the expected values.  These data were obtained
during the precision evaluation. Based on this precision data, a standard deviation was calculated and the
MDLs were defined as 3 times the standard deviation for each target analyte.  The precision-based MDLs
were calculated for soil samples that had been homogenized, dried, ground, and the intrusive-prepared level
placed in a sample cup.  The precision-based MDLs for the X-MET 920-MP are shown in Table 4-3. The
precision-based MDLs for all analytes except chromium were obtained using a 120-second count time for
the Cd109 source.  The chromium MDL was determined using a 180-second count time. The developer did
not provide specific MDLs for the analytes in Table 4-3; however, it did indicate the MDLs should be
similar to those for the X-MET 920-P with the Si(Li) detector.
                                               49

-------
    Another method of determining MDLs involved the direct comparison of the FPXRF and reference
data. When these sets of data are plotted against each other, the results were linear. As the plotted line
approached zero, there was a point at which the slope of the curve changes.  Figure 4-2 illustrates this
effect for arsenic.  This point was determined by observation and was somewhat subjective; however, an
analysis showed that even a 25 percent error in identifying this point resulted in only a 10 percent change in
MDL.  By determining the values of this FPXRF concentration and multiplying it by two standard
deviations around the mean, it was  possible to determine a field or performance-based MDL for the
analyzer. This result is also shown in Figure 4-3.
 Table 4-3. Method Detection Limits
Precision-based Field-based
Analyte MDL (mg/kg) MDL (mg/kg)
Arsenic
Barium
Chromium
Copper
Lead
Nickel
Zinc
50
330
115
50
30
Not determined
30
175
Not determined
130
Not determined
105
Not determined
210
  Note:   mg/kg  Milligrams per kilogram.
                                                   "
                                                   3
                                                   ro
                                                   Q
                                                    ro
                                                   <
                                                   o
                                                   CM
                                                   O)
                                                      10000
     1000
      100
                                                        10
                                                          10          100         1000
                                                                   Reference Data (mg/kg)
                                                                                           10000
Figure 4-2.  Critical Zone for the Determination
of a Field-based Method Detection Limit for
Copper: Between 100 and 200 mg/kg for the
reference data, the linear relationship between the
two data sets change. This point of change was
used to determine the field-based MDL.
    As would be expected, the precision-based MDLs generally improved when the absorption energy of
the analyte was close to the excitation energy of the Cd109 source. Detection limits could not be determined
for nickel because the nickel concentrations were too low.  A field-based MDL for barium and copper
could not be determined because the regression plots were linear through the y-intercept at zero. The field-
based MDLs for the remaining three analytes were 3 to 7 times higher than the precision-based MDLs.
Detection limits for chromium were consistent using either technique.

Throughput

    Throughput for the X-MET 920-MP varied during the demonstration since differing count times were
used. The count times varied from 30 seconds at the ASARCO site to 180 seconds at the RV Hopkins site.
The following discussion includes operator time only and takes into consideration the time spent for daily
set up,  QC checks, and for data handling. The throughput does not take into consideration sample
preparation time.

    At the ASARCO site, 30-second count times were used for all in situ -unprepared and in situ -prepared
analyses.  On the first day of in situ -unprepared analysis, the operator was only able to achieve a
throughput of three samples per hour due to problems with computer batteries and software lockouts. By
the third day of in s/YM-unprepared analyses, a throughput of eight samples per hour was achieved. When
                                               50

-------
the work moved indoors at the ASARCO site and the X-MET 920-MP was run from alternating current
power, the operator was able to analyze 14 in s/YH-prepared samples per hour.

   At the RV Hopkins site, the count times were increased to 60 seconds for the in s/YH-unprepared and
prepared samples for better chromium resolution and quantitation. Despite the increase in count time,
throughput for the in situ -unprepared and in situ -prepared samples was 8  and 14 samples per hour,
respectively.

   Using a 60-second count time for the intrusive-unprepared samples at ASARCO, the operator analyzed
176 samples in a 12.5-hour day for a throughput of 14 samples per hour.  This was the maximum number
of samples analyzed in one day.  When the count time was increased to  120 seconds  and  180 seconds for
the intrusive-prepared samples at the two sites, the throughput dropped to 11  and 8 samples per hour,
respectively. Metorex claimed that 200 - 400 samples could be analyzed in an 8- to  10-hour day based on
10- to 100-second count times.  The throughput during this demonstration averaged between 70 and 150
samples per day, depending on count times.

Drift

   Drift is a measurement of an analyzer's variability in quantitating a known amount of a standard over
time.  Drift was evaluated by reviewing results from the periodic analysis of a mid-level SSCS from each
site.

   Metorex recommended that the operator run the SSCS as a check sample once per hour during the
demonstration. The ASARCO SSCS check sample was  analyzed 57 times and the RV Hopkins SSCS
check sample 42 times during the demonstration. These  data were reduced to RSDs for the target analytes
and the percent drift from the mean recovery of the true value. The percent drift from the mean recovery
was averaged for each day and is shown in Figure 4-3  for the target analytes reported by the X-MET 920-
MP.
on

0
"c 0
0)
CL
-^n


D
	 ri 	
-° B D
1 B | 1 g 1 1
B n
-n n
n n
n
Barium Arsenic Chromium Copper Lead Zinc Nickel
Analyte






  Figure 4-3. Drift Summary:  This graph shows the general drift of the analyzer's results in
  measuring a check sample. Each point represents a different day's analysis of the same sample.
  The daily fluctuations exhibited for each analyte are a direct representation of drift.
                                              51

-------
    The X-MET 920-MP displayed the least drift for arsenic and lead with RSD values less than 3.6
percent for both SSCS check samples. Greater drift was shown for barium, chromium, copper, nickel, and
zinc. The drift for barium, chromium, and nickel was only assessed using the RV Hopkins SSCS check
sample.  Barium and nickel showed the greatest drift as supported by RSD values of 19.5 and 16.2,
respectively. Chromium had an RSD value of 6.7.  Copper and zinc had RSD values of 9.1 and 5.0,
respectively, for the ASARCO SSCS check sample, but RSD values of 23.5 and 24.9, respectively, for the
RV Hopkins SSCS check sample.  This may indicate a difference in the empirical calibration between the
two sites, and specifically, in defining the overlap tables.  All data points for barium, copper, and zinc in
Figure 4-3 outside of 10 percent drift were for the RV Hopkins SSCS check sample.

Intramethod Assessment

    Intramethod measures of each analyzer's performance included results on analyzer blanks,
completeness, precision, and accuracy.  The following narrative discusses these characteristics.

Blanks

    Analyzer blanks for the X-MET 920-MP consisted of pure lithium carbonate placed directly in a
sample cup and lithium carbonate that had been put through all four sample preparation steps like the dried
and ground soil samples. The blanks were used to monitor for contamination of the probe by material such
as residual soil left on the probe face. Six blanks were analyzed at the ASARCO site using the ASARCO
empirical calibration. Four blanks were analyzed at the RV Hopkins site using the RV Hopkins empirical
calibration.

    The X-MET 920-MP reported values for arsenic and copper above the precision-based MDLs for the
ASARCO blanks and values for chromium, lead, and zinc above the  precision-based MDLs for the RV
Hopkins blanks. None of the other analytes were detected in the blanks. An examination of the data
showed the results for the analytes found above their respective MDLs were an artifact of the regression
equations used for calibration.  It was apparent that the regression equations developed using SSCSs did
not apply to a much different matrix such as lithium carbonate. The blank data for the X-MET 920-MP
demonstrated that a "clean" matrix matching the SSCSs should be used for a blank.  It was therefore
concluded that the lithium carbonate blanks did not show  a contamination problem.

Completeness

    The X-MET 920-MP produced data for 1,168 out of the 1,260 samples for a completeness of 92.7
percent, slightly below the demonstration objective of 95 percent.  The 92 samples for which no data were
obtained were all from the ASARCO site.  There were 78 in 5/YM-unprepared, one in s/YH-prepared, and 13
intrusive-prepared samples for which no data were obtained. The loss of data for these samples was a
combination of software problems, operator oversight, and a failure of the lead chemist to realize the data
set was not complete prior to returning the rental computer to the  manufacturer.  It should be noted that
none of the lost data was caused by mechanical or electronic malfunctions of the analyzer.

Prec/'s/OA7

    Precision was expressed in terms of the percent RSD between replicate measurements. The percent
RSD is defined as the SD divided by the mean concentration times 100. The precision data for the target
analytes detectable by the analyzer are shown in Table 4-4.  The precision data reflected in the range of 5

                                              52

-------
to 10 times the MDL reflects the precision generally referred to in analytical methods such as SW-846 and
represents general method precision.

    The analyzer performed 10 replicate measurements on 12 soil samples that had analyte concentrations
ranging from less than 50 mg/kg to tens of thousands of milligrams per kilogram.  Each of the 12 soil
samples underwent the four sample preparation steps. Therefore, there was a total of 48 precision points
for the analyzers. The replicate measurements were taken using the same  source count times used for
regular sample analysis.  For each detectable analyte  in each precision sample, a mean concentration, SD,
and RSD were calculated.

    In this demonstration, the analyzer's precision or RSD for a given analyte had to be less than or equal
to 20 percent to be considered quantitative screening level data and less than or equal to 10 percent to be
considered definitive level data.  The analyzer's precision data, reflected by its precision data in the 5 to 10
times MDL range, one measure in the overall data quality evaluation, were all below the 10 percent RSD
required for definitive level data quality classification. Chromium and nickel did not have sufficient data to
allow data quality conclusions based on precision. The lower precision for nickel  in the 50 - 500 mg/kg
range may be an artifact of the low concentrations of nickel in the soil samples and the influence of iron
interference in the samples.

             Table 4-4.  Precision Summary
                                Mean % RSD Values by Concentration Range
             Analyte
 5-10 Times
MDLa (mg/kg)
50 - 500
(mg/kg)
500-1,000
 (mg/kg)
 >1,000
(mg/kg)
Arsenic
Barium
Chromium
Copper
Lead
Nickel
Zinc
1.91 (16)
5.91 (4)
ND
7.92(12)
7.60(16)
ND
2.10(20)
3.08 (20)
ND
8.19(16)
6.81 (16)
4.16(16)
20.92 (4)
3.06 (36)
1 .32 (8)
9.52 (8)
ND
15.93(4)
6.85(12)
ND
1 .00 (4)
0.73 (4)
10.08(8)
ND
4.19(8)
5.43(12)
ND
ND
             Notes:          The MDLs referred to in this column are the precision-based MDLs
                            shown in Table 4-3.
                     mg/kg  Milligrams per kilogram.
                        ND  No data.
                         ( )  Number of samples, including the four sample preparation steps,
                            each consisting of 10 replicate analyses.

    The precision data in Table 4-4 shows there was little effect of concentration on the precision. The
precision samples were purposely chosen to span a large concentration range to test the effect of analyte
concentration on precision. It was expected that as the analyte concentration increased, the precision would
increase. However, the precision was so good even for the low concentration samples, that little or no
increase in precision was seen as the concentrations increased.  With the exception of nickel, all mean RSD
values were less than 10 percent for samples containing less than  500 mg/kg of an analyte. The good
precision for this instrument can be attributed to the high efficiency of the gas proportional detector and to
the manner in which the spectrum is deconvoluted by the software.
                                                53

-------
    Table 4-5 shows the precision data as affected by sample preparation and count time. It was not
possible to separate the two factors as both were changing throughout the demonstration. For the analytes
barium, chromium, nickel, and zinc, there was no consistent improvement in precision with increased
sample preparation and count time.  There was a slight increase in arsenic precision with increased sample
preparation and count time.  The most dramatic increase in precision was seen for copper and lead,
especially for the intrusive samples analyzed at longer count times.  A paired t-test was used to determine if
the average RSD values were significantly different for copper and lead for the different sample
preparation techniques and count times. A significance level of 0.05 was used to test for a difference in
mean RSD values. For copper, the precision for the intrusive samples was significantly different (better)
than for the in situ samples. There was no statistical difference for the copper precision between the
intrusive-prepared and intrusive-unprepared samples. The precision for lead was found to be statistically
equal for the in situ -unprepared and in situ -prepared samples. The precision for lead for the intrusive
samples was found to be statistically significantly better than for the in situ samples.

 Table 4-5. Precision as Affected by Sample Preparation and Count Times
 Analyte
                                             Mean % RSD Values
  In Situ-Unprepared   In Situ-Prepared   Intrusive-Unprepared   Intrusive-Prepared
    (30 Seconds)        (30 Seconds)         (60 Seconds)         (120 Seconds)
Arsenic
Barium3
Chromium3
Copper
Lead
Nickel3
Zinc
3.45
8.17
6.77
13.91
11.74
17.45
2.24
3.11
11.15
8.04
14.62
7.32
24.65
2.72
2.29
6.77
9.06
11.51
1.84
ND
2.57
1.69
12.16
8.87
7.44
1.81
20.67
3.59
 Notes:
    RSD values obtained for RV Hopkins precision samples only. Count times were 60 seconds for
    all in situ samples, 120 seconds for intrusive-unprepared, and 180 seconds for intrusive-
    prepared samples.
ND  No data.
Accuracy
    Accuracy refers to the degree to which a measured value for a sample agrees with a reference or true
value for the same sample. Originally, it was intended that accuracy would be assessed for the X-MET
920-MP using six site-specific PE samples and 14 SRMs.  However, since the X-MET 920-MP was
calibrated empirically using SSCSs whose true values were determined using EPA SW-846 Methods
3050A/6010A, the analysis of the SRMs would not be a true assessment of accuracy since the matrix
(particle size, contaminant analytes, and interferant analytes) of the SRMs did not match the matrix on
which the calibration was based. Also, the true values in the SRMs were determined using "total metals"
methods unlike the  SSCSs, which had analyte concentrations determined using EPA SW-846 Methods
3050A/6010A, a partial digestion method.

    The site-specific PE samples consisted of three samples from each of the two demonstration sites that
were collected during the predemonstration activities and sent to six independent laboratories for analysis
by laboratory-grade XRF analyzers. The mean measurement for each analyte was used as the true value
                                               54

-------
concentration.  Because the PE samples matched the matrix at each site and had true values determined by
a method closely matching FPXRF, these PE samples would serve as a good accuracy check. However,
because the X-MET 920-MP used Methods 3050A/6010A-based data for calibration, the analytical
method used to determine the PE sample's concentrations did not match the analytical method used to
calibrate the analyzer.  Therefore, the site-specific PEs could not be used as an ideal assessment of
accuracy.

    Ultimately, it was concluded that none  of the site-specific PEs or SRMs could be used as a true
accuracy check for this analyzer. However, during this demonstration, the  site-specific PEs were submitted
to the reference laboratory for analysis by EPA SW-846 Methods 3050A/6010A. It was decided that the
reference laboratory data would be used as  the true value for the site-specific PEs. However, only one
analysis was performed on each site-specific PE sample and that there is a degree of error associated with
any single measurement.  The only alternative would have been to send the site-specific PE samples to
multiple laboratories for analysis by EPA SW-846 Methods 3050A/6010A and then determine the mean
analyte concentrations and the standard error around those means. This was not considered prior to the
demonstration and, therefore, was considered an option.

    Although not a perfect assessment of accuracy, the reference laboratory data was used as the true
values for the site-specific PE samples.  The PEs did not have published acceptance  ranges.  As specified in
the demonstration plan, an acceptance range of 80 - 120 percent recovery of the true value was used to
evaluate accuracy for the six site-specific PEs. Table 4-6 summarizes the accuracy  data for the target
analytes for the X-MET 920-MP. Figure 4-4 shows the true value, the measured value, and percent
recovery for the individual PEs. No figure  is presented for the nickel data because there were only two
detects.

  Table 4-6. Accuracy Summary for Site-Specific PE Results

 Analyte
         Percent
         Within        Mean     Range of     SD of
      Acceptance     Percent     Percent     Percent
         Range      Recovery   Recovery   Recovery
Concentration
Range (mg/kg)
Site-Specific Performance Evaluation Samples
Arsenic
Barium
Chromium
Copper
Nickel
Lead
Zinc
4
4
3
6
3
6
6
0
25
0
50
0
0
50
57
151
46
143
469
64
74
44-68
88 - 226
22-69
113-209
369 - 569
52-79
38- 102
12.3
59.5
23.4
38.0
141
11.3
21.3
105-20,885
195-3,817
139- 1,761
69 - 5,869
129-321
242- 13,210
116-3,497
  Notes:
    n Number of samples with detectable analyte concentrations.
   SD Standard deviation.
mg/kg Milligrams per kilogram.
                                              55

-------
                         Arsenic

           I Measured Value  CDTrue Value
                                             40
                                    • Percent Recovery
                                                       10000
                                                                                           250
                   Barium

    • Measured Value  CUTrue Value
                                                                                  • Percent Recovery
      10000
       1000
    o
    O
        100
                                                        10000
                                                         1000
                                                          100
                                                                                            240
                                                      O
                                             20
                        Chromium
          • Measured Value  CDTrue Value
                                   • Percent Recovery
                    Copper
     • Measured Value CUTrue Value
                                                                                    • Percent Recovery
      100000
       10000
    g   1000
    o
    O
        100
         10
                                             100
                                             80
                                             60
                                             40
                                                        10000
                                                         1000
                                                0
                                                QL
o
O
                                                          100
                                                           10
120

100

80

60

40

20
                          Lead

          • Measured Value  CUTrue Value
                                   • Percent Recovery
                     Zinc

     • Measured Value CUTrue Value
                                                                                    • Percent Recovery
  Figure 4-4. Site-specific PE Sample Results: These graphs illustrate the relationship between the
  analyzer's data (measured values) and the true values for the site-specific PE samples.  The gray
  bars represent the percent recovery for the analyzers.  Each set of three bars (black, white, and gray)
  represents a single site-specific PE sample.

    Based on the 80 - 120 percent recovery acceptance criteria, the analyzer produced 7 out of 32 results
or 21.9 percent within the acceptance range for all analytes in the six PEs.  The X-MET 920-MP produced
0 percent of the results within the acceptance range for arsenic, chromium, nickel, and lead; 25 percent for
barium; and 50 percent for copper and zinc. The mean percent recovery and range of percent recoveries
demonstrate that the X-MET 920-MP results were biased high for barium, copper, and nickel. The X-
MET 920-MP nickel results showed the most disparity from the true values, possibly due to interferences
from iron in these samples.  The copper and zinc X-MET 920-MP results showed the closest comparison to
the true values. The X-MET 920-MP underestimated arsenic, chromium, and lead. In comparing the
results of the check sample analysis discussed earlier in the drift assessment section, it is interesting to note
the mean percent recoveries for the target analytes  in the check sample were similar to the mean percent
recoveries listed in Table 4-6. This indicates that the low or high biases for specific analytes were
consistent throughout the demonstration.
                                                 56

-------
Intermethod Assessment

    The comparison of the X-MET 920-MP results to those of the reference methods was performed using
the statistical methods detailed in Section 2.  The purpose of this evaluation was to determine the
comparability between data produced by the analyzer and that produced by the reference laboratory. If the
Iog10 transformed FPXRF data were statistically equivalent to the Iog10 transformed reference data and had
acceptable precision (10 percent RSD or less), the data met the definitive level criteria. If the data did not
meet the definitive level criteria, but could be mathematically corrected to be equivalent to the reference
data, they met the quantitative screening level criteria. If the analyzer did not meet the definitive level
criteria, and the statistical evaluation could not identify a predictable bias in the data but the analyzer
identified the presence or absence of contamination with at least a 90 percent accuracy rate, the data was
classified as qualitative screening level.

    The analyzer was configured to report concentrations for all of the target analytes except iron,
cadmium, and antimony. Due to a variety of technical problems, some analyzer data were considered not
valid and subsequently dropped from the comparability assessment.

    The comparability assessment begins with a regression analysis of the entire data set for the primary
analytes, arsenic and lead which had r2 values above 0.85.  Barium, chromium, copper, and zinc had r2
values ranging from 0.69 to 0.55. Based on the slope of the regressions, the analyzer tended to
overestimate barium and underestimate arsenic, copper, zinc, and chromium concentrations relative to the
reference methods. (The slope values were determined with the FPXRF data plotted on the y-axis and the
reference data plotted on the x-axis.) The r2 value for nickel, 0.32, showed poor comparability; however,
the number of data points was limited.

    The next step in the data evaluation  involved assessing the potential impact of the site, soil type, and
sample preparation variables on the regression analysis (Table 4-7).  The effect of the site variable was
assessed for lead and zinc only.  These were the only target analytes exhibiting a wide and similar
concentration distribution at each site. The evaluation of the soil type variable was also limited. Copper
and arsenic did not exhibit a wide concentration distribution in the clay soil, 10 - 250 mg/kg and 4-28
mg/kg, respectively.  Barium, chromium, and nickel did not exhibit a wide concentration distribution in
either the sand or loam soil, 30 - 300 mg/kg (95 percent of data), 10 - 130 mg/kg, and 8 - 550 mg/kg,
respectively. Based on this evaluation, there was no apparent impact of either the site or soil type variables
on the regression.

    The sample preparation variable was confounded by the increased count times used for each successive
sample preparation step. For this reason, any increase in comparability cannot be wholly attributed to
either sample preparation or count times. The precision study indicated that only copper and lead data
showed increases in precision with increased count times. The lead exhibited a 10-fold increase in
precision over the four sample preparation-increased count time steps, and copper exhibited a twofold
increase in precision over the same  intervals.  Based on this, for all target analytes, except lead and copper,
any increase in comparability with increased sample preparation was attributed to sample preparation.
Changes in lead and copper comparability remain confounded, including both sample preparation and count
time effects.
                                                57

-------
Table 4-7. Regression Parameters3 by Primary Variable
               Arsenic
                                                               Barium
n r2 Std. Err. Y-lnt. Slopeb ^^^jj||||j^^J n f2 std E|T Y-lnt. Slopeb
746
746
ND
368
383
ND
135
208
209
197
0.946
0.946
ND
0.926
0.957
ND
0.898
0.969
0.964
0.937
0.14
0.14
ND
0.17
0.11
ND
0.21
0.10
0.11
0.15
0.49
0.49
ND
0.63
0.42
ND
0.62
0.42
0.44
0.47
0.80
0.80
ND
0.74
0.82
ND
0.75
0.82
0.81
0.80
All Data
ASARCO Site
RV Hopkins Site
Sand Soil
Loam Soil
Clay Soil
In Situ-Unprepared
In Situ-Prepared
Intrusive-Unprepared
Intrusive-Prepared
113
ND
113
ND
ND
116
ND
ND
41
74
0.620
ND
0.550
ND
ND
0.550
ND
ND
0.580
0.660
0.47
ND
0.57
ND
ND
0.57
ND
ND
0.51
0.46
0.37
ND
0.15
ND
ND
0.15
ND
ND
0.13
0.20
1.08
ND
1.15
ND
ND
1.15
ND
ND
1.14
1.16
  928
             Chromium
         r2   Std. Err.  Y-lnt. Slope"
                                           Variable
                                                               Copper
                                                               Std. Err.  Y-lnt. Slope"
0.680
0.23
0.20
0.82
      All Data
747
0.690
0.46
 0.17
0.88
  709
0.006
0.06
1.27
0.04
   ASARCO Site
618
0.730
0.43
-0.38
1.06
  219
0.650
0.30
0.61
0.69
  RV Hopkins Site
129
0.390
0.35
 0.59
0.81
  348
0.001
0.02
1.30
0.01
     Sand Soil
321
0.761
0.22
 0.19
0.82
  385
0.003
0.20
1.30
0.08
     Loam Soil
288
0.660
0.58
-0.86
1.12
  187
0.650
0.30
0.61
0.69
     Clay Soil
129
0.390
0.35
 0.59
0.81
  124
0.180
0.04
1.17
0.11
 In Situ-Unprepared
106
0.550
0.34
 0.20
0.82
  214
0.260
0.04
1.09
0.16
  In Situ-Prepared
142
0.700
0.36
 0.24
0.76
  298
0.793
0.18
0.18
0.80
Intrusive-Unprepared
233
0.734
0.36
 0.40
0.74
  282
0.639
0.29
0.43
0.76
 Intrusive-Prepared
184
0.716
0.29
 0.66
0.73
n r2 Std. Err. Y-lnt. Slopeb ^^^jj||||j^^J n f2 std E|T Y-lnt. Slopeb
849
700
149
340
363
146
127
193
275
251
0.878
0.885
0.760
0.910
0.841
0.760
0.850
0.920
0.840
0.890
0.25
0.24
0.27
0.21
0.26
0.27
0.30
0.20
0.27
0.23
-0.06
-0.19
0.46
-0.26
-0.04
0.46
-0.12
-0.23
0.04
-0.05
1.00
1.05
0.82
1.06
1.01
0.82
1.01
1.07
0.96
1.00
All Data
ASARCO Site
RV Hopkins Site
Sand Soil
Loam Soil
Clay Soil
In Situ-Unprepared
In Situ-Prepared
Intrusive-Unprepared
Intrusive-Prepared
130
ND
130
ND
ND
130
ND
ND
47
79
0.320
ND
0.320
ND
ND
0.320
ND
ND
0.250
0.470
0.47
ND
0.47
ND
ND
0.47
ND
ND
0.42
0.33
0.17
ND
0.17
ND
ND
0.17
ND
ND
0.54
0.24
0.79
ND
0.79
ND
ND
0.79
ND
ND
0.58
0.77
                                             58

-------
Table 4-7. Continued
          r2    Std. Err.  Y-lnt.  Slope"
                                             Variable
  667
  619
    48
  271
   354
    49
   101
   179
   196
   192
0.550
0.527
0.730
0.498
0.537
0.730
0.480
0.640
0.662
0.417
0.36
0.34
0.34
0.44
0.25
0.34
0.39
0.30
0.36
0.39
 0.29
 0.39
 0.40
 0.13
 0.71
 0.40
 0.10
 0.40
-0.09
 0.56
0.85
0.79
0.98
0.84
0.70
0.98
0.87
0.81
0.99
0.78
      All Data
   ASARCO Site
  RV Hopkins Site
     Sand Soil
     Loam Soil
      Clay Soil
 In Situ-Unprepared
  In Situ-Prepared
Intrusive-Unprepared
 Intrusive-Prepared
Notes:        a Regression parameters based on Iog10 transformed data. These parameters were determined
               using the FPXRF data as the dependent variable.  Therefore, these parameters cannot be used
               to correct FPXRF data. See Section 5.
             b Slope values determined with FPXRF data plotted on the y-axis and the reference data plotted on
               the x-axis.
             n Number of data points.
         Y-lnt. Y-lntercept.
       Std. Err. Standard error.
           ND Analytes not present in significant quantities to provide meaningful regression. This data
               evaluation does not include copper data from ASARCO samples 102 to 201 for intrusive-
               prepared analyses and RV Hopkins data from both in situ sample preparation analyses.
    The sample preparation variable exhibited the greatest impact on the regression analysis. Generally,
the largest shift in the r2 was exhibited between the in s/YM-unprepared and in s/YH-prepared analyses
(Figure 4-5).  Except for chromium, sample homogenization and increased count times accounted for
between 80 and 100 percent of the total increase in the r2 experienced across all sample preparation steps.
Chromium exhibited its greatest increase in comparability between the last in situ and first intrusive
analysis.

    The impact of the soil texture and site variables was also assessed for each of the four sample
preparation steps (Tables 4-8 and 4-9). This evaluation was conducted for lead and zinc only.  These were
the only primary analytes exhibiting relatively even concentration distribution between the site and soil
variables.  No clear effect on comparability was observed for these variables. A minor trend was observed
for the site variable.  Generally, the r2 values for lead at the ASARCO site were higher, relative to the RV
Hopkins site. This may have been due to the wider range of analyte  concentrations found at the ASARCO
site and not a site or soil related effect.
                                                59

-------
"3 100000
^)
E
| 10000
Q-
o 1000
LJJ
X 10°
1C
In situ-unprepared

*##
^t+++ +
M$I^++




0 1000 10000 100000
Reference Data (mg/kg)

-55 100000
^>
E
•2 10000
Q
Q.
° 1000
LJJ
^ 100
1C
Intrusive-unprepared

• ^^
w^^,



)0 1000 10000 100000
Reference Data (mg/kg)










-5; 100000
^)
E
~ 10000
Q
Q-
o 1000
LJJ
x 100
1C
In situ-prepared

^^
- ^^v
^^,
)0 1000 10000 100000
Reference Data (mg/kg)

-3 100000
«
D)
E
| 10000
Q-
o 1000
LJJ
2
X 10°
1C
Intrusive-prepared

J&-
i H=kip
H^f^,
)0 1000 10000 100000
Reference Data (mg/kg)
  Figure 4-5. Sample Preparation Effects on Arsenic Results: These log-log plots illustrate the
  change in comparability with changes in sample preparation step.
    Within the sample preparation steps, the effect of contaminant concentration was also examined.  The
data sets for the primary analytes were sorted into the following concentrations ranges: 0-100 mg/kg, 100
- 1,000 mg/kg, and greater than 1,000 m/kg.  The regression analysis for each target analyte and for each
sample preparation step was rerun on these concentration-sorted data sets.  A review of these results
showed general improvement in the r2 and standard error for each target analyte with increasing
concentration.  The 0-100 mg/kg concentration range showed the poorest comparability.  This is most
likely due to this range generally occurring just at or below the analyzers MDLs.  The analyzer's precision
and accuracy are lowest in this concentration range. Generally, the r2 values improved among the  100 -
1,000 mg/kg and greater than 1,000 mg/kg ranges. This data indicated that there was a concentration
effect on comparability. This effect appears to be linked to the general proximity of a measurement to its
associated MDL. The further away from the MDL, the less effect concentration will have on quantitation
and comparability.
                                               60

-------
Table 4-8. Regression Parameters3 for the Preparation Variable Sorted by Soil Texture
               Arsenic
                                                               Barium
n r2 Std. Err. Y- Slopeb ^^^^^^^^1 n r2 Std. Err. Y-lnt. Slopeb
In Situ-Unprepared
78
57
ND
0.904
0.892
ND
0.20
0.21
ND
0.67
0.54
ND
0.72
0.78
ND
In Situ-Prepared
95
112
ND
0.946
0.981
ND
0.15
0.07
ND
0.56
0.39
ND
0.77
0.84
ND
Intrusive-Unprepared
97
114
ND
0.928
0.973
ND
0.17
0.08
ND
0.61
0.42
ND
0.75
0.82
ND
Intrusive-Prepared
98
101
ND
0.926
0.942
ND
0.17
0.13
ND
0.65
0.30
ND
0.75
0.86
ND
Soil Texture
Sand Soil
Loam Soil
Clay Soil
Soil Texture
Sand Soil
Loam Soil
Clay Soil
Soil Texture
Sand Soil
Loam Soil
Clay Soil
Soil Texture
Sand Soil
Loam Soil
Clay Soil
In Situ-Unprepared
ND
ND
44
ND
ND
0.574
ND
ND
0.48
ND
ND
0.46
ND
ND
1.05
In Situ-Prepared
ND
ND
29
ND
ND
0.702
ND
ND
0.29
ND
ND
0.77
ND
ND
0.91
Intrusive-Unprepared
ND
ND
39
ND
ND
0.482
ND
ND
0.49
ND
ND
0.65
ND
ND
0.97
Intrusive-Prepared
ND
ND
74
ND
ND
0.660
ND
ND
0.46
ND
ND
0.20
ND
ND
1.16
              Chromium
              Std. Err.   Y-    Slopeb
                         Int.
          In Situ-Unprepared
                                  Soil Texture
                                                               Copper
                                                               Std. Err. Y-lnt.  Slopeb
                                                  In Situ-Unprepared
   69
0.108
0.03
1.19
 0.09
Sand Soil
67   0.646    0.24
                 0.07  0.88
   52
0.000
0.03
1.33
 0.00
Loam Soil
38
0.523
0.44
-0.84
1.14
   96   0.450    0.31
                 1.27
                0.46
                   Clay Soil
                            68   0.236    0.53
                                       0.80  0.80
           In Situ-Prepared
                                  Soil Texture
                                                   In Situ-Prepared
   96
0.000
0.02
1.31
 0.00
Sand Soil
80
0.638
0.22
 0.68  0.59
  114
0.520
0.03
1.02
 0.21
Loam Soil
61
0.715
0.45
-0.42
0.95
   97   0.365    0.25
                 1.57
                0.30
                   Clay Soil
                            68   0.084    0.40
                                       1.55  0.33
         Intrusive-Unprepared
                                  Soil Texture
                                                 Intrusive-Unprepared
   98
0.012
0.02
1.33
-0.02
Sand Soil
93
0.895
0.17
 0.13  0.83
  113
0.361
0.03
1.11
 0.15
Loam Soil
85
0.828
0.33
-0.84
1.09
   93   0.633    0.30
                 0.64
                0.66
                   Clay Soil
                            51   0.439    0.36
                                       0.25  0.96
          Intrusive-Prepared
                                  Soil Texture
                                                  Intrusive-Prepared
   85
0.000
0.01
1.31
 0.00
Sand Soil
81
0.929
0.13
-0.56  1.17
  102
   94
0.003
0.673
0.23
0.31
1.80
0.59
-0.07
 0.71
Loam Soil
Clay Soil
25
77
0.951
0.413
0.12
0.31
-0.92
 0.75
1.26
0.74
                                             61

-------
Table 4-8. Continued
n r2 Std. Err. Y- Slopeb ^^^^^^^^1 n r2 Std. Err. Y-lnt. Slopeb
In Situ-Unprepared
72
53
40
0.885
0.840
0.500
0.26
0.32
0.47
-0.17
-0.11
-0.63
1.02
1.02
1.12
In Situ-Prepared
86
106
35
0.941
0.875
0.577
0.17
0.22
0.30
-0.29
0.04
0.03
1.08
0.99
0.91
Intrusive-Unprepared
90
108
74
0.920
0.880
0.667
0.20
0.22
0.29
-0.36
-0.10
0.86
1.10
1.02
0.69
Intrusive-Prepared
86
94
72
0.938
0.813
0.856
0.17
0.29
0.24
-0.18
-0.05
0.07
1.05
1.04
0.94
Soil Texture
Sand Soil
Loam Soil
Clay Soil
Soil Texture
Sand Soil
Loam Soil
Clay Soil
Soil Texture
Sand Soil
Loam Soil
Clay Soil
Soil Texture
Sand Soil
Loam Soil
Clay Soil

In Situ-Unprepared
ND
ND
23
ND
ND
0.217
ND
ND
0.83
ND
ND
-2.31
ND
ND
2.04
In Situ-Prepared
ND
ND
33
ND
ND
0.325
ND
ND
0.68
ND
ND
-0.75
ND
ND
1.08
Intrusive-Unprepared
ND
ND
45
ND
ND
0.248
ND
ND
0.41
ND
ND
0.50
ND
ND
0.60
Intrusive-Prepared
ND
ND
79
ND
ND
0.470
ND
ND
0.33
ND
ND
0.24
ND
ND
0.77

n r' Std. Err. Y- Slope ^^^^
In Situ-Unprepared
51
50
91
0.558
0.506
0.620
0.35
0.36
0.19
-0.11
0.27
1.09
0.86
0.87
0.54
In Situ-Prepared
75
106
92
0.664
0.566
0.607
0.40
0.25
0.20
0.03
0.70
1.08
0.91
0.71
0.52
Intrusive-Unprepared
71
107
23
0.641
0.630
0.712
0.46
0.25
0.43
-0.26
0.36
-0.54
0.98
0.82
1.31
Intrusive-Prepared
72
91
31
0.007
0.452
0.816
0.42
0.17
0.34
1.72
1.37
0.29
0.12
0.42
1.09
Soil Texture
Sand Soil
Loam Soil
Clay Soil
Soil Texture
Sand Soil
Loam Soil
Clay Soil
Soil Texture
Sand Soil
Loam Soil
Clay Soil
Soil Texture
Sand Soil
Loam Soil
Clay Soil
Notes:             Regression parameters based on Iog10 transformed data. These parameters were determined using the FPXRF
                  data as the dependent variable. Therefore, these parameters cannot be used to correct FPXRF data. See Section
                  5.
                b Slope values determined with FPXRF data plotted on the y-axis and reference data plotted on the x-axis.
            Y-lnt. Y-lntercept.
          Std. Err. Standard Error.
                n Number of data points.
              ND Analytes not present in significant quantities to provide meaningful regression. This data does not include copper
                  data from ASARCO samples 102 to 201 for the intrusive-prepared analyses.
                                                          62

-------
Table 4-9.  Regression Parameters3 for the Preparation Variable Sorted by Site Name
Arsenic ^^^^^^H Barium
n r Std. Err. Y- Slope" ^^^^^^^B n r Std. Err. Y-lnt. Slope"
In Situ-Unprepared
135
ND
0.898
ND
0.21
ND
0.62
ND
0.75
ND
In Situ-Prepared
208
ND
0.969
ND
0.10
ND
0.42
ND
0.82
ND
Intrusive-Unprepared
209
ND
0.964
ND
0.11
ND
0.44
ND
0.81
ND
Intrusive-Prepared
197
ND
0.937
ND
0.15
ND
0.47
ND
0.80
ND
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins
In Situ-Unprepared
ND
44
ND
0.574
ND
0.48
ND
0.46
ND
1.05
In Situ-Prepared
ND
29
ND
0.702
ND
0.29
ND
0.77
ND
0.91
Intrusive-Unprepared
ND
39
ND
0.482
ND
0.49
ND
0.65
ND
0.97
Intrusive-Prepared
ND
74
ND
0.660
ND
0.46
ND
0.20
ND
1.16
Chromium ^^^^^^H Copper
n r2 Std. Err. Y- Slope" ^^^^^^^B n r2 Std. Err. Y-lnt. Slope"
In Situ-Unprepared
121
96
0.069
0.450
0.03
0.31
1.22
1.27
0.07
0.46
In Situ-Prepared
209
97
0.227
0.365
0.03
0.25
1.14
1.57
0.12
0.30
Intrusive-Unprepared
211
93
0.156
0.633
0.03
0.30
1.18
0.64
0.09
0.66
Intrusive-Prepared
188
94
0.003
0.673
0.26
0.31
1.38
0.59
0.10
0.71
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins
In Situ-Unprepared
104
68
0.621
0.236
0.31
0.53
0.10
0.80
0.86
0.80
In Situ-Prepared
140
68
0.738
0.084
0.33
0.40
0.20
1.55
0.77
0.33
Intrusive-Unprepared
180
51
0.822
0.439
0.29
0.36
-0.15
0.25
0.90
0.96
Intrusive-Prepared
107
77
0.923
0.413
0.15
0.31
-0.64
0.75
1.19
0.74
                                           63

-------
Table 4-9. Continued
Lead ^^^^^^H Nickel
n r Std. Err. Y- Slope" ^^^^^^^B n r2 Std. Err. Y-lnt. Slope"
In Situ-Unprepared
126
40
0.859
0.500
0.29
0.47
-0.17
-0.63
1.03
1.12
In Situ-Prepared
193
35
0.915
0.577
0.20
0.30
-0.23
0.03
1.07
0.91
Intrusive-Unprepared
201
74
0.881
0.667
0.24
0.29
-0.25
0.86
1.06
0.69
Intrusive-Prepared
181
72
0.878
0.856
0.24
0.24
-0.17
0.07
1.06
0.94
^^^^^^^^^^^^^^^^^^^^^^^1
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins


n r2 Std. Err. Y- Slope ^^^H
In Situ-Unprepared
104
91
0.459
0.620
0.43
0.19
-0.01
1.09
0.91
0.54
In Situ-Prepared
179
92
0.640
0.607
0.30
0.20
0.40
1.08
0.81
0.52
Intrusive-Unprepared
174
23
0.665
0.712
0.32
0.43
0.11
-0.54
0.90
1.31
Intrusive-Prepared
161
31
0.229
0.816
0.30
0.34
1.21
0.29
0.46
1.09
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins
Site Name
ASARCO
RV Hopkins
In Situ-Unprepared
ND ND ND ND ND
23 0.217 0.83 -2.31 2.04
In Situ-Prepared
ND ND ND ND ND
34 0.346 0.78 -1.26 1.28
Intrusive-Unprepared
ND ND ND ND ND
45 0.248 0.41 0.50 0.60
Intrusive-Prepared
ND ND ND ND ND
79 0.470 0.33 0.24 0.77



Notes:         a Regression parameters based on Iog10 transformed data.  These parameters were determined using the FPXRF data
                 as the dependent variable.  Therefore, these parameters cannot be used to correct FPXRF data.  See Section 5.
               b Slope values determined with FPXRF data plotted on the y-axis and reference data plotted on the x-axis.
           Y-lnt. Y-lntercept.
        Std. Err. Standard Error.
               n Number of data points.
             ND Analytes not present in significant quantities to provide meaningful regression. This data does not include copper
                 data from ASARCO samples 102 to 201 for the intrusive-prepared analyses.
                                                          64

-------
    Another way to examine the comparability between the two methods involves measuring the average
relative bias and accuracy between the FPXRF data and the reference. These measurements were made by
using the raw FPXRF and reference data.  The average relative bias indicates the average factor by which
the two data sets differ. Concentration effects can affect bias. For example, it is possible for an analyzer
to greatly underestimate low concentrations but greatly overestimate high concentrations and have a
relative bias of zero.  To eliminate this concentration effect, the data can be corrected by a regression
approach (see Section 5).  Otherwise, only narrow concentration ranges can be analyzed, or average
relative accuracy can be analyzed. The average relative accuracy is the average factor by which each
individual analyzer measurement differs from the corresponding reference measurement.

    A final decision regarding the assignment of data quality levels involves an assessment of both the r2
and precision results. Using the criteria presented in Table 2-2, a summary of the X-MET analyzer's
performance during this demonstration is provided in Table 4-10.

  Table 4-10.  Summary of Data Quality Level Parameters
X-MET Precision Mean Method Detection Coefficient of
Target 920-MP % RSD (mg/kg) Limits (mg/kg) Determination Data Quality
Analytes Analytes 5 - 10 X MDL (Precision-based) (r2 all Data) Level
Arsenic
Barium
Chromiu
m
Copper
Lead
Zinc
Nickel
Iron
Cadmium
Antimony
Arsenic
Barium
Chromium
Copper
Lead
Zinc
Nickel
Not Reported
Not Reported
Not Reported
1.901
5.91
ND
7.92
7.60
2.10
ND
-
-
-
50
330
115
50
30
30
ND
-
-
-
0.946
0.620
0.680
0.690
0.878
0.550
0.320
-
-
-
Definitive
Qualitative Screening
Insufficient Data
Qualitative Screening
Definitive
Qualitative Screening
Insufficient Data
-
-
-
                                               65

-------
                                          Section  5
                 Applications Assessment and Considerations
    The X-MET 920-MP is designed to provide qualitative and quantitative data on the concentration of
metals in soils, sludges, and other solids. X-MET software uses empirical site-specific calibration and
quantitation to maximize instrument performance and account for common soil-related matrix
interferences. This analyzer is designed for field use; however, its physical and power requirements made it
cumbersome to use in the in situ mode. The analyzer's design is best suited for benchtop applications. The
analyzer experienced some software operating problems resulting in downtime and lost data during the 1-
month field demonstration. During this time, more than 1,260 samples were measured by the analyzer.
The training provided by the developer was sufficient to allow basic field operation of the analyzer;
however, frequent developer assistance was required to address software problems throughout the
demonstration. The complexity of an empirical calibration suggested that the analyzer's calibration and
performance may improve with increased operator familiarity. The developer provided highly accessible
and timely field support. Much of the downtime occurred while using the analyzer in the in situ mode. The
data lost was a result of software problems and involved less than 8 percent of the data.

    Comparison of the analyzer's Iog10 transformed data to the Iog10 transformed reference data indicated
that the analyzer could produce definitive level quality data for arsenic and lead. This indicated that the
analyzer's data were statistically equivalent to the reference data for these  analytes. For the other target
analytes, barium, copper, and zinc, the analyzer produced qualitative screening level data. The X-MET
920-MP exhibited instrument precision similar to the reference methods, indicating high measurement
reproducibility.

    The analyzer's probe uses one radioactive source with count times ranging from 60 to 180 live-
seconds.  Longer count times generally increase accuracy and lower the detection limits but decrease
sample throughput.  The throughput for the analyzer ranged from 8 to 14 samples per hour. A summary of
key operational features is listed in Table 5-1.

    There were no apparent effects of site or soil type on performance.  This demonstration identified
sample preparation as the most important variable with regard to comparability of the FPXRF analyzer to
the reference method.  For copper and lead, this preparation effect was confounded by changing count
times. Therefore, increased comparability for lead and copper, with increased preparation, may also be the
result of the associated increase in count  times.  The analyzer can be applied in an in situ  or intrusive
mode. The data from this demonstration indicated that when operated in the in situ mode, the user most
probably would not be able to show a strong correlation between FPXRF and reference data.  This may not
be due to instrument error, but rather to inherent spacial variability of contamination, even within an area
as small as the 4-inch by 4-inch grid sampled during this demonstration. The greatest increase in


                                               66

-------
correlation between the FPXRF data and reference data for the analyzer was achieved after the initial
sample preparation step (sample homogenization). Further sample preparation, such as sieving or drying
and sieving, in most cases did improve the comparability; however, relative to the demonstration-defined
data quality criteria, no increases in data quality level occurred.

 Table 5-1.  Summary of Test Results and Operational Features
 Field portable—Total weight of 10 pounds for probe and electronics unit
 Sample throughput (8 to 14 samples per hour)
 Produces qualitative screening level data for barium, copper, and zinc, arsenic, and lead at the
 definitive level
 Data is linearly related to EPA SW-846 Methods 3050A/6010A data
 Precision—Percent RSD values less than 10 percent at 5 to 10 times the MDL for all analytes
 Generally not susceptible to soil matrix effects
 Can be used on soils exhibiting up to 30 percent water saturation by weight
 Operation requires detailed training (1 to 3 days)
 The need for an auxiliary computer reduces the sample throughput in the in situ mode.  Battery
 lifetime of only 2 hours on the auxiliary computer
 Empirical calibration requires 10 to 20 well characterized and homogenized site-specific samples
 A single source limits the number of elements that can be reported
    Based on this demonstration, the analyzer is well suited to the real-time assessment of metals
contamination in soil samples. This demonstration showed that the use of this analyzer in an in situ mode
is less efficient than applying it in an intrusive mode. In addition, more extensive training or operator
experience was needed to reduce the potential for data loss and downtime associated with software and
calibration problems. Although in several cases the analyzer produced data statistically equivalent to the
reference data, generally confirmatory analysis is required for FPXRF analysis. If 10 - 20 percent of the
samples measured by the analyzer are submitted for reference method analysis, instrument bias relative to
standard methods such as 3050A/6010A can be determined.  This will only hold true if the analyzer and the
reference laboratory measure similar samples. This was accomplished in this demonstration by thorough
sample homogenization. The demonstration showed that the analyzer exhibits a strong linear relationship
with the reference data more than a 5 orders of magnitude concentration range. For optimum correlation
and bias correction, samples with high, medium, and low concentration ranges from  a project must be
submitted for reference  method analysis.  Table 5-2 shows the effects of data correction for the in situ-
prepared data set. Changes in average relative bias and accuracy are used to show the effects of data
correction for the in situ -prepared samples.

    The steps to correct the FPXRF measurements to more closely match the reference data are as follows:
1.   Conduct sampling and FPXRF analysis.
2.   Select 10 - 20 percent of the  sampling locations for resampling. These  resampling locations can be
    evenly distributed over the range of concentrations measured or they can focus on an action level
    concentration range.
3.   Resample the selected locations. Thoroughly homogenize the samples and have  each sample analyzed
    by FPXRF and a reference method.


                                               67

-------
Table 5-2.  Effects of Data Correction on FPXRF Comparability to Reference Data for All In Situ-
           Prepared Samples
                    Average
                  Relative Bias
Target Analyte    on Raw Data3
 Average
 Relative
 Bias on
Corrected
  Datab
  Average
  Relative
  Average
  Relative
Accuracy on
Accuracy on    Corrected
 Raw Data0       Datad
 Acceptable
  Relative
  Accuracy
Based on PE
  Samples6
Arsenic
Barium
Chromium
Copper
Lead
Nickel
Zinc
0.89
4.14
1.01
1.90
0.87
0.46
1.03
1.03
1.19
1.24
2.62
1.07
1.20
1.21
1.35
4.00
1.92
1.63
1.39
2.72
1.90
1.27
1.74
1.98
1.97
1.46
2.43
1.90
1.76
1.36
1.55
1.18
1.63
1.56
1.64
Notes     A measurement of average relative bias, measured as a factor by which the FPXRF, on
:          average, over- or underestimates results relative to the reference methods. This
          measurement of bias is based on raw (not Iog10 transformed) data. This average relative
          bias does not account for any concentration effect on analyzer performance.
         b
          A measurement of average relative bias on the FPXRF data after it has been corrected
          using the eight-step regression approach.

          A measurement of average relative accuracy at the 95 percent confidence interval,
          measured as a factor by which the raw FPXRF, on average, over- or underestimates
          individual results relative to the reference methods. This measurement of accuracy is
          based on raw (not Iog10 transformed) data.  This average relative accuracy is independent
          of concentration  effects.
         d
          A measurement of average relative accuracy at the 95 percent confidence interval, of the
          corrected FPXRF data obtained using the eight-step regression approach.

          A measurement of accuracy represents a factor and 95 percent confidence interval that
          define the acceptable range of differences allowed between the reference method
          reported concentrations and the true value concentrations in the PE samples.  This bias is
          included only as  a general reference for assessing the improvement on comparability of
          FPXRF data and reference data  after FPXRF data correction.

The average relative bias is  calculated as follows:

   Average relative bias = ((£j[FPXRFj/Referencej])/number of paired samples)-1

This value represents the percentage that the FPXRF over- or underestimates the reference data, on
average, for the entire data set.  To convert this calculated value to a factor, 1.0 is added to the
calculated average relative bias.  The above table presents the average  relative bias as a factor.

The average relative accuracy is calculated as follows:

   Average relative accuracy =SQRT (£|([FPXRFj/Referencej]-1)2/number of paired sample)

This value represents the percentage that an individual FPXRF measurement over- or underestimates
the reference data. The relative accuracy numbers in the table are calculated at the 95 percent
confidence interval. This is accomplished by adding two standard deviations to the above formula
before the square root is taken.  To convert this calculated value to a factor,  1.0 is added to the
calculated average relative accuracy. The above table presents the average relative bias as a factor.
                                            68

-------
4.  Tabulate the resulting data with reference data in the x-axis column (independent variable) and the
    FPXRF data in the y-axis column (dependent variable).  Transform this data to the equivalent Iog10
    value for each concentration.
5.  Conduct a linear regression analysis and determine the r2, y-intercept, and slope of the relationship.
    The r2 should be greater than 0.70 to proceed.
6.  Place the regression parameters into Equation 5-1:
        7(log10 corrected FPXRF data)  =  slope *(log10 FPXRF data) +  Y-intercept                    (5-1)

7.  Use the above equation with the Iog10  transformed FPXRF results from Step 4 above and calculate the
    equivalent Iog10 corrected FPXRF data.
8.  Take the anti-log10 (10 [i°gio transformed corrected FPXRF data]) of the equivalent Iog10 corrected FPXRF data
    calculated in Step 7.  These resulting values (in milligrams per kilogram) represent the corrected
    FPXRF data.

    To show the effect of correcting the FPXRF data, the change in average relative bias and accuracy can
be examined.  The average relative bias between the FPXRF data and the reference data is a measure of the
degree to which the FPXRF over- or underestimates concentrations relative to the reference methods.  The
relative bias is an average number for the  entire data set and may not be representative of individual
measurements. An example of this can be seen in an analyzer's data where measurements are
underestimated at low concentrations but overestimated at high concentrations.  On average, the relative
bias for this analyzer is zero; however, this bias is not representative for high or low concentration
measurements. To avoid this dilemma, three approaches can be taken:  (1) the evaluation of average
relative bias can be focused on a narrow concentration range, (2) the analyzer's data can be corrected using
the regression approach described above,  or (3) average relative accuracy can be calculated. Average
relative accuracy represents the percentage that an individual measurement is over- or underestimated
relative to a reference measurement.  Table 5-2 shows the average relative bias and accuracy exhibited by
the FPXRF, before and after data correction using the eight-step approach previously discussed.

    The average relative bias and accuracy for the analytes falling into the definitive level data quality
category were generally small.  However, the analytes falling into the quantitative and qualitative  screening
level data quality categories had generally larger average relative bias and accuracy.

    In cases where the regression derived corrected average relative accuracy is worse than the raw average
relative accuracy, such as seen in Table 5-2 for chromium, copper, lead, nickel, and zinc, the eight-step
FPXRF data correction approach presented earlier may not be appropriate. For these elements, data at or
below the field-based MDLs were responsible for the decreased accuracy.  If the data set in question is
representative of the entire population of data being characterized, then the raw FPXRF data can  simply be
multiplied by the raw average relative accuracy factor for correction. However, the eight-step regression
approach should be used whenever the performance of the analyzer is strongly concentration-dependent or
if the  sample population being used for data correction is not representative of the entire data population
being characterized.
                                                69

-------
    The Metorex X-MET 920-MP Analyzer can provide a rapid assessment of the distribution of metals
contamination at a hazardous waste site. This data can be used to characterize general site contamination,
guide critical conventional sampling and analysis, and monitor removal actions. This demonstration
suggested that in some applications and for some analytes, the FPXRF data may be statistically similar to
the reference data.  The creation of draft Method 6200 will help in the acceptance of FPXRF data for all
definitive level applications and possibly qualitative screening level applications.  The FPXRF data can be
produced and interpreted in the field on a daily or per sample  basis. This real-time analysis allows the use
of contingency-based sampling for any application and greatly increases the potential for meeting project
objectives on a single mobilization.

General Operational Guidance

    The following paragraphs describe general operating considerations for FPXRF analysis.  This
information is based on the SW-846 Method 6200 for FPXRF analysis.

    General operation of FPXRF instruments will vary according to specific developer protocols.  For all
environmental applications, confirmatory or reference sampling should be conducted so that FPXRF data
can be corrected. Before operating any FPXRF instrument, the developer's manual should be consulted.
Most developers recommend that their instruments be allowed to warm up for 15 - 30 minutes before
analysis of samples. This will help alleviate drift or energy calibration problems.

    An FPXRF instrument should be operated according to the developer's recommendations. There are
two modes in which FPXRF instruments can be operated:  in situ and intrusive. The in situ mode involves
analysis of an undisturbed soil or sediment sample. Intrusive  analysis involves collecting and preparing a
soil or sediment sample before analysis. Some FPXRF instruments can operate in both modes of analysis,
while others are designed to operate in only one mode. The two modes of analysis are discussed below.

    For in situ analysis, one requirement is that any large or nonrepresentative debris be removed from the
soil surface before analysis.  This debris includes rocks, pebbles, leaves, vegetation, roots, and concrete.
Another requirement is that the soil surface be as smooth as possible so that the probe window will have
good contact with the surface. This may require some leveling of the  surface with a stainless-steel  trowel.
Many developers recommend that the soil be tamped down to increase soil density and compactness.  This
step reduces the influence of soil density variability on the results.  During the demonstration, this modest
amount of sample preparation was found to take less than 5 minutes per sample location.  The last
requirement is that the soil or sediment not be saturated with water. Most FPXRF instruments will perform
adequately for soils with moisture contents of 5 - 20 percent, but will not perform well for saturated soils,
especially if ponded water exists on the surface. Data from this demonstration did not indicate an effect on
data quality from soil moisture content. Source count times for in situ analysis usually range from 30 to
120 seconds, but source count times will vary between instruments and on required detection  limits.

    For intrusive analysis of surface soil or sediment, it is recommended that a sample be collected from a
4- by 4-inch square that is 1  inch deep. This will produce a soil sample of approximately 375 grams or
250 cm3, which is  enough soil to fill an 8-ounce jar. The sample should be homogenized, dried, and ground
before analysis. The data from this demonstration indicated that sample preparation, beyond
homogenization, does not greatly improve data quality. Homogenization can be conducted by kneading a
soil sample in a plastic bag.  One way to monitor homogenization when the sample is kneaded in a plastic
bag is to add sodium fluorescein dye to the sample. After the  moist sample has been homogenized, it is
examined under an ultraviolet light to assess the distribution of sodium fluorescein throughout the sample.

                                               70

-------
If the fluorescent dye is evenly distributed in the sample, homogenization is considered complete; if the dye
is not evenly distributed, mixing should continue until the sample has been thoroughly homogenized.
During the demonstration, the homogenization procedure using the fluorescein dye required 3 to 5 minutes
per sample.

    Once the soil or sediment sample has been homogenized, it can be dried. This can be accomplished
with a toaster oven or convection oven. A small portion of the sample (20 - 50 grams) is placed in a
suitable container for drying. The sample should be dried for 2 to 4 hours in the convection or toaster oven
at a temperature not greater than 150 °C. Microwave drying is not recommended.  Field studies have
shown that microwave drying can increase variability between the  FPXRF data and reference data.  Fligh
levels of metals in a sample can cause arcing in the microwave oven, and sometimes slag will form in the
sample. Microwave oven drying can also melt plastic containers used to hold the sample.

    The homogenized, dried sample material can also be ground with a mortar and pestle and passed
through a 60-mesh sieve to achieve a uniform particle size.  Sample grinding should continue until at least
90 percent of the original sample passes through the sieve. The grinding step normally averages 10
minutes per sample.

    After a sample is prepared, a portion of the sample should then be placed in a 31-mm polyethylene
sample cup (or equivalent) for analysis. The sample cup should be completely filled and covered with a
2.5-micrometer Mylar™ (or equivalent) film for analysis. The rest of the soil sample should be placed in a
jar, labeled, and archived.  All equipment, including the mortar, pestle, and sieves, must be thoroughly
cleaned so that the  sample method blanks are below the MDLs of the procedure.
                                               71

-------
                                        Section 6
                                       References
Havlick, Larry L., and Ronald D. Grain. 1988. Practical Statistics for the Physical Sciences. American
    Chemical Society. Washington, B.C.

Kane, J. S., S. A. Wilson, J. Lipinski, and L. Butler. 1993. "Leaching Procedures: A Brief Review of
    Their Varied Uses and Their Application to Selected Standard Reference Materials." American
    Environmental Laboratory.  June.  Pages 14-15.

Kleinbaum, D. G., and L. L. Kupper.  1978. Applied Regression Analysis and Other Multivariable
    Methods.  Wadsworth Publishing Company, Inc., Belmont, California.

Morgan, Lewis,  & Bockius.  1993. RODScan®.

PRC Environmental Management, Inc.  1995.  "Final Demonstration Plan for Field Portable X-ray
    Fluorescence Analyzers."

U.S. Environmental Protection Agency.  1993.  "Data Quality Objectives Process for Superfund-Interim
    Final Guidance."  Office of Solid Waste and Emergency Response. Washington, D.C.
    EPA/540/R-93/071.
                                             72

-------