METHOD 300.1

   DETERMINATION OF INORGANIC ANIONS IN DRINKING WATER BY ION
                         CHROMATOGRAPHY
                              Revision 1.0
John D. Pfaff (USEPA, ORD, NERL) - Method 300.0, (1993)

Daniel P. Hautman (USEPA, Office of Water) and David J. Munch (USEPA, Office of Water)
Method 300.1, (1997)
              NATIONAL EXPOSURE RESEARCH LABORATORY
                OFFICE OF RESEARCH AND DEVELOPMENT
               U.S. ENVIRONMENTAL PROTECTION AGENCY
                        CINCINNATI, OHIO 45268
                                 300.1-1

-------

-------
                                   METHOD 300.1

        DETERMINATION OF INORGANIC ANIONS IN DRINKING WATER
                           BY ION CHROMATOGRAPHY
1.    SCOPE AND APPLICATION

     1.1   This method covers the determination of the following inorganic anions in reagent
           water, surface water, ground water, and finished drinking water. As a result of
           different specified injection volumes (See conditions in Tables 1A and IB), these
           anions are divided between the common anions listed in Part A and the inorganic
           disinfection by-products listed in Part B.  These different injection volumes are
           required in order to compensate for the relative concentrations of these anions in
           drinking water and maintain good chromatographic peak shape throughout the
           expected dynamic range of the detector.  Bromide is included in both Part A, due to
           its importance as a common anion, as well as Part B due to its critical role as a
           disinfection by-product precursor.

           PART A.-- Common Anions
           Bromide              Nitrite
           Chloride              ortho-Phosphate-P
           Fluoride               Sulfate
           Nitrate

           PART B.— Inorganic Disinfection By-products
           Bromate              Chlorite
           Bromide              Chlorate

     1.2   The single laboratory Method Detection Limits (MDL, defined in Sect. 3.11) for the
           above analytes are listed in Tables 1 A, IB and 1C.  The MDL for a specific matrix
           may differ from those listed, depending upon the nature of the sample and the specific
           instrumentation employed.

           1.2.1   In order to achieve comparable detection limits, an ion chromatographic
                  system must utilize suppressed conductivity detection, be properly
                  maintained and must be capable of yielding a baseline with no more than 5 nS
                  noise/drift per minute of monitored response over the background
                  conductivity.

     1.3   This method is recommended for use only by or under the supervision of analysts
           experienced in the use of ion chromatography and in the interpretation of the
           resulting ion chromatograms.
                                        300.1-2

-------
      1.4  When this method is used to analyze unfamiliar samples for any of the above anions,
           anion identification should be supported by the use of a fortified sample matrix
           covering the anions of interest.  The fortification procedure is described in Sect.
           9.4.1.

      1.5  Users of the method data should state the data-quality objectives prior to analysis.
           Users of the method must demonstrate the ability to generate acceptable results with
           this method, using the procedures described in Sect. 9.0.

      1.6  Bromide and nitrite react with most oxidants employed as disinfectants. The utility
           of measuring these anions in treated water should be considered prior to conducting
           the analysis.

2.    SUMMARY OF METHOD

      2.1   A small volume of sample, 10 ^L for Part A and 50 fiL for Part B, is introduced into
           an ion chromatograph. The anions of interest are separated and measured, using a
           system comprised of a guard column, analytical column, suppressor device, and
           conductivity detector.

      2.2  The ONLY difference between Parts A and B is the volume of sample analyzed by
           the ion chromatographic system. The separator columns and guard columns as well
           as eluent conditions are identical.

3.    DEFINITIONS

      3.1   ANALYSIS BATCH - A group of no more than 20 field samples (Field sample
           analyses include only those samples derived from a field sample matrix.  These
           include the initial and duplicate field samples as well as all Laboratory Fortified
           Sample Matrices). The analysis batch must include an Initial Calibration Check
           Standard, an End Calibration Check Standard, Laboratory Reagent Blank, and a
           Laboratory Fortified Blank. Within an ANALYSIS BATCH, for every group often
           field samples, at least one Laboratory Fortified Matrix (LFM) and either a Field
           Duplicate, a Laboratory Duplicate or a duplicate of the LFM must be analyzed.
           When more than 10 field samples are analyzed, a Continuing Calibration Check
           Standard must be analyzed after the tenth field sample analysis.

      3.2   CALIBRATION STANDARD (CAL) ~ A solution prepared from the primary
           dilution standard solution or stock standard solutions and the surrogate analyte. The
           CAL solutions are used to calibrate the instrument response with respect to analyte
           concentration.
                                        300.1-3

-------
      3.2.1  INITIAL CALIBRATION STANDARDS ~  A series of CAL solutions used
            to initially establish instrument calibration and develop calibration curves for
            individual target anions.

      3.2.2  INITIAL CALIBRATION CHECK STANDARD - An individual CAL
            solution, analyzed initially, prior to any sample analysis, which verifies
            previously established calibration curves.

      3.2.3  CONTINUING CALIBRATION CHECK STANDARD - An individual
            CAL solution which is analyzed after every tenth field sample analyses which
            verifies the previously established calibration curves and confirms accurate
            analyte quantitation for the previous ten field samples analyzed.

      3.2.4  END CALIBRATION CHECK STANDARD - An individual CAL solution
            which is analyzed after the last field sample analyses which verifies the
            previously established calibration curves and confirms accurate analyte
            quantitation for all field samples analyzed since the last continuing calibration
            check.

3.3   FIELD DUPLICATES ~ Two separate samples collected at the same time and place
      under identical circumstances and treated exactly the same throughout field and
      laboratory procedures. Analyses of field duplicates indicate the precision associated
      with sample collection, preservation and storage, as well as with laboratory
      procedures.

3.4   INSTRUMENT PERFORMANCE CHECK SOLUTION (IPC) - A solution of one
      or more method analytes, surrogates, or other test substances used to evaluate the
      performance of the instrument system with respect to a defined set of criteria.

3.5   LABORATORY DUPLICATE - Two sample aliquots, taken in the laboratory from
      a single sample bottle, and analyzed separately with identical procedures.  Analyses
      of LD1 and LD2 indicate precision associated specifically with the laboratory
      procedures, removing any associated variables attributed by sample collection,
      preservation, or storage procedures.

3.6   LABORATORY FORTIFIED BLANK (LFB) - An aliquot of reagent water or
      other blank matrices to which known quantities of the method analytes are added in
      the laboratory. The LFB is analyzed exactly like a sample, and its purpose is to
      determine whether the methodology is in control, and whether the laboratory is
      capable of making accurate and precise measurements.

3.7   LABORATORY FORTIFIED SAMPLE MATRIX (LFM) - An aliquot of an
      environmental sample to which known quantities of the method analytes are added in
                                 300.1-4

-------
      the laboratory. The LFM is analyzed exactly like a sample, and its purpose is to
      determine whether the sample matrix contributes bias to the analytical results.  The
      background concentrations of the analytes in the sample matrix must be determined
      in a separate aliquot and the measured values in the LFM corrected for background
      concentrations.

 3.8   LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or other
      blank matrices that are treated exactly as a sample including exposure to all
      glassware, equipment, solvents, reagents, and surrogates that are used with other
      samples.  The LRB is used to determine if method analytes or other interferences are
      present in the  laboratory environment, the reagents, or the apparatus.

 3.9   LINEAR CALIBRATION RANGE (LCR) ~ The concentration range over which
      the instrument response is linear.

 3.10  MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided by
      vendors concerning a chemical's toxicity, health hazards, physical properties, fire, and
      reactivity data including storage, spill, and handling precautions.

 3.11  METHOD DETECTION LIMIT (MDL) -- The minimum concentration of an
      analyte that can be identified, measured and reported with 99% confidence that the
      analyte concentration is greater than zero.

 3.12  MINIMUM REPORTING LEVEL (MRL) -- The minimum concentration that can
      be reported for an anion in a sample following analysis. This defined concentration
      can be no lower than the concentration of the lowest calibration standard and can
      only be used if acceptable quality control criteria for this standard are met.

 3.13  PERFORMANCE EVALUATION SAMPLE (PE) -- A certified solution of method
      analytes whose concentration is unknown to the analyst. Often, an aliquot of this
      solution is added to a known volume of reagent water and analyzed with procedures
      used for samples. Results of analyses are used to determine statistically the accuracy
      and precision that can be expected when a method is performed by a competent
      analyst.

3.14  QUALITY CONTROL SAMPLE (QCS) ~ A solution of method analytes of known
      concentrations that is used to fortify an aliquot of LRB or sample matrix.  The QCS
      is obtained from a source external to the laboratory and different from the source of
      calibration standards. It is used to check laboratory performance with externally
      prepared test materials.

3.15  SURROGATE ANALYTE -- An analyte added to a sample, which is unlikely to be
      found in any sample at significant concentration, and which is added directly to a
                                  300.1-5

-------
           sample aliquot in known amounts before any sample processing procedures are
           conducted.  It is measured with the same procedures used to measure other sample
           components. The purpose of the surrogate analyte is to monitor method
           performance with each sample.

      3.16 STOCK STANDARD SOLUTION (SSS) ~ A concentrated solution containing one
           or more method analytes prepared in'the laboratory using assayed reference materials
           or purchased from a reputable commercial source.

4.    INTERFERENCES

      4.1   Interferences can be divided into three different categories:  direct chromatographic
           coelution, where an analyte response is observed at very nearly the same retention
           time as the target anion; concentration dependant coelution, which is observed when
           the response of higher than typical concentrations of the neighboring peak overlap
           into the retention window of the target anion; and,  ionic character displacement,
           where retention times may significantly shift due to the influence of high ionic
           strength matrices (high mineral content or hardness) overloading the exchange sites
           in the column and significantly shortening target analyte's retention times.

           4.1.1    A direct chromatographic coelution may be solved by changing columns,
                   eluent strength, modifying the eluent with organic solvents (if compatible
                   with 1C columns), changing the detection systems, or selective removal of
                  the interference with pretreatment. Sample dilution will have little to no
                   effect.

           4.1.2   Sample dilution may resolve some of the difficulties if the interference is the
                   result of either concentration dependant coelution or ionic character
                   displacement, but it must be clarified that sample dilution will alter your
                   Minimum Reporting  Limit (MRL) by a proportion  equivalent to that of
                  the dilution. Therefore, careful consideration of project objectives should be
                   given prior to performing such a dilution. An alternative to sample dilution,
                   may be dilution of the eluent as outlined in 11.9.

           4.1.3    Pretreatment cartridges can be effective as a means to eliminate certain
                   matrix interferences. Prior to using any pretreatment, the analyst should be
                   aware that all instrument calibration standards must be pretreated in exactly
                  the same manner as the pretreated unknown field samples. The need for
                  these cartridges have been greatly reduced with recent advances in high
                   capacity anion exchange columns.

                  4.1.3.1     Extreme caution should be exercised in using these pretreatment
                             cartridges.  Artifacts are known to leach from certain cartridges
                                         300.1-6

-------
                       which can foul the guard and analytical columns causing loss of
                       column capacity indicated by shortened retention times and
                       irreproducible results. Frequently compare your calibration
                       standard chromatograms to those of the column test
                       chromatogram (received when the column was purchased) to
                       insure proper separation and similar response ratios between the
                       target analytes is observed.

4.2   Method interferences may be caused by contaminants in the reagent water, reagents,
      glassware, and other sample processing apparatus that lead to discrete artifacts or
      elevated baselines in an ion chromatogram. These interferences can lead to false
      positive results for target analytes as well as reduced detection limits as a
      consequence of elevated baseline noise.

4.3   Samples that contain particles larger than 0.45 microns and reagent solutions that
      contain particles larger than 0.20 microns require filtration to prevent damage to
      instrument columns and flow systems.

4.4   Any anion that is only weakly retained by the column may elute in the retention time
      window of fluoride and potentially interfere. At concentrations of fluoride above 1.5
      mg/L, this interference may not be significant, however, it is the responsibility of the
      user to generate precision and accuracy information in each sample matrix.

4.5   Close attention should be given to the potential for carry over peaks from one
      analysis which will effect the proper detection of analytes of interest in a second,
      subsequent analysis. Normally, the elution of sulfate (retention time of 13.8 min.)
      indicates the end of a chromatographic run, but, in the ozonated and chlorine dioxide
      matrices, which were included as part of the single operator accuracy and bias study
      (See Table 2B), a small response (200 nS baseline rise) was observed for a very late
      eluting unknown peak at approximately 23 minutes. Consequently, a run time of 25
      minutes is recommended to allow for the proper elution of any potentially interferant
      late peaks.  It is the responsibility of the user to confirm that no late eluting peaks
      have carried over into a subsequent analysis thereby compromising the integrity of
      the analytical results.

4.6   Any residual chlorine dioxide present in the sample will result in the formation of
      additional chlorite prior to analysis. If any concentration of chlorine dioxide is
      suspected in the sample, the sample must be purged with an inert gas (helium, argon
      or nitrogen) for approximately five minutes or until no chlorine dioxide remains.  This
      sparging must be conducted prior to ethylenediamine preservation and at time of
      sample collection.
                                   300.1-7

-------
5.    SAFETY

      5.1    The toxicity or carcinogenicity of each reagent used in this method have not been
            fully established. Each chemical should be regarded as a potential health hazard and
            exposure should be as low as reasonably achievable.  Cautions are included for
            known extremely hazardous materials or procedures.

      5.2    Each laboratory is responsible for maintaining a current awareness file of OSHA
            regulations regarding the safe handling of the chemicals specified in this method. A
            reference file of Material Safety Data Sheets (MSDS) should be made available to all
            personnel involved in the chemical analysis. The preparation of a formal safety plan
            is also advisable.

      5.3    The following chemicals have the potential to be highly toxic or hazardous, consult
            MSDS.

            5.3.1    Sulfuric acid — When used to prepared a 25 mN sulfuric acid regenerant
                   solution for chemical suppression using a Dionex Anion Micro Membrane
                   Suppressor (AMMS).

6.    EQUIPMENT AND SUPPLIES

      6.1    Ion chromatograph — Analytical system complete with ion chromatograph and all
            required accessories including syringes, analytical columns, compressed gasses and a
            conductivity detector.

            6.1.1    Anion guard column: Dionex AG9-HC, 2 mm (P/N 52248), or equivalent.
                   This column functions as a protector of the separator column.  If omitted
                   from the system the retention times will be shorter.

            6.1.2    Anion separator column: Dionex AS9-HC column, 2 mm (P/N 52244), or
                   equivalent. The microbore (2 mm) was selected in the development of this
                   method as a means to tighten the bromate elution band and thus reduce the
                   detection limit. An optional column (2 mm or 4 mm) may be used if
                   comparable resolution of peaks is obtained, and the requirements of Sect. 9.0
                   can be met.  The AS9-HC, 2 mm column using the conditions outlined in
                   Table 1A and IB produced the separation shown in Figures 1  through 4.

                   6.1.2.1    If a 4 mm column is employed, the injection volume should be
                            raised by a factor of four to 40 uL for Part A anions and 200  |^L
                            for Part B anions in order to attain comparable detection limits. A
                            four fold increase in injection volume compensates  for the four
                                        300.1-8

-------
                        fold increase in cross sectional surface area of the 4 mm standard
                        bore column over the 2 mm microbore column.

              6.1.2.2    Comparable results can be attained using the Dionex, AS9-HC, 4
                        mm column. MDLs for the part B, inorganic disinfection by-
                        products using this 4 mm column are displayed along with analysis
                        conditions in Table 1C.

       6.1.3   Anion suppressor device: The data presented in this method were generated
              using a Dionex Anion Self Regenerating Suppressor (ASRS, P/N 43187).
              An equivalent suppressor device may be utilized provided comparable
              detection limits are achieved and adequate baseline stability is attained as
              measured by a combined baseline drift/noise of no more than 5 nS per minute
              over the background conductivity.

              6.1.3.1    The ASRS was set to perform electrolytic suppression at a current
                        setting of 100 mA using an external source DI water mode.
                        Insufficient baseline stability was observed using the ASRS in
                        recycle mode.

      6.1.4   Detector -- Conductivity cell (Dionex  CD20, or equivalent) capable of
              providing data as required in Sect. 9.2.

6.2   The Dionex Peaknet Data Chromatography Software was used to generate all the
      data in the attached tables.  Systems using a strip chart recorder and integrator or
      other computer based data system may achieve approximately the same MDL's but
      the user should demonstrate this by the procedure outlined in Sect. 9.2.

6.3   Analytical balance, ±0.1 mg sensitivity.  Used to accurately weigh target analyte salts
      for stock standard preparation.

6.4   Top loading balance, ±10 mg sensitivity. Used to accurately weigh reagents to
      prepare eluents.

6.5   Weigh boats,  plastic, disposable - for weighing eluent reagents.

6.6   Syringes, plastic, disposable, 10 mL - used during sample preparation.

6.7   Pipets, Pasteur, plastic or glass, disposable, graduated, 5 mL and 10 mL.

6.8   Bottles, high density polyethylene (HOPE), opaque or glass, amber, 30 mL, 125 mL,
      250 mL.  For sampling and storage of calibration solutions.  Opaque or amber due to
      the photoreactivity of chlorite anion.
                                   300.1-9

-------
      6.9   Micro beakers, plastic, disposable - used during sample preparation.

7.    REAGENTS AND STANDARDS

      7.1   Reagent water:  Distilled or deionized water, free of the anions of interest. Water
           should contain particles no larger than 0.20 microns.

      7.2   Eluent solution : Sodium carbonate (CASRN 497-19-8) 9.0 mM. Dissolve 1.91 g
           sodium carbonate (NajCOj) in reagent water and dilute to 2 L.

           7.2.1   This eluent solution must be purged for 10 minutes with helium prior to use
,                  to remove dissolved gases which may form micro bubbles in the 1C
                  compromising system performance and adversely effecting the integrity of
                  the data.

      7.3   Stock standard solutions, 1000 mg/L (1  mg/mL):  Stock standard solutions may be
           purchased as certified solutions or prepared from ACS reagent grade, potassium or
           sodium salts as listed below, for most analytes. Chlorite requires careful
          , consideration as outline below in 7.3.5.1.

           7.3.1   Bromide (Br) 1000 mg/L:  Dissolve 0.1288 g sodium bromide (NaBr,
                  CASRN 7647-15-6) in reagent water and dilute to 100 mL in a volumetric
                  flask.

           7.3.2   Bromate (BrO3") 1000 mg/L:  Dissolve 0.1180 g of sodium bromate
                  (NaBrO3, CASRN 7789-38-0) in reagent water  and dilute to 100 mL in a
                  volumetric flask.

           7.3.3   Chlorate (C103') 1000 mg/L: Dissolve 0.1275 g of sodium chlorate
                  (NaC103, CASRN 7775-09-9) in reagent water  and dilute to 100 mL in a
                  volumetric flask.

           7.3.4   Chloride (Cl') 1000 mg/L: Dissolve 0.1649 g sodium chloride (NaCl,
                  CASRN 7647-14-5) in reagent water and dilute .to 100 mL in a volumetric
                  flask.

           7.3.5   Chlorite (C102') 1000 mg/L: Assuming an exact 80.0 % NaC102 is
                  amperometrically titrated from technical grade NaC102(See Sect. 7.3.5.1).
                  Dissolve 0.1676 g of sodium chlorite (NaC102, CASRN 7758-19-2) in
                  reagent water and dilute to 100 mL in a volumetric flask.

                  7.3.5.1    High purity sodium chlorite (NaClO2) is not currently
                            commercially available due to potential  explosive instability.


                                       300.1-10

-------
                 Recrystallization of the technical grade (approx. 80%) can be
                 performed but it is labor intensive and time consuming. The
                 simplest approach is to determine the exact % NaCIO 2 using the
                 iodometric titration procedure (Standard Methods, 19th Ed.,
                 4500-C1O2.C).  Following titration, an individual component
                 standard of chlorite must be analyzed to determine if there is any
                 significant contamination (greater than 1% of the chlorite weight)
                 in the technical grade chlorite standard from any of the Part B
                 components. These contaminants will place a high bias on the
                 calibration of the other anions if all four Part B components are
                 mixed in an combined calibration solution. If these other anions
                 are present as contaminants, a separate chlorite calibration needs
                 to be performed.

7.3.6   Fluoride (F) 1000 mg/L:  Dissolve 0.2210 g sodium fluoride (NaF, CASRN
       7681-49-4) in reagent water and dilute to 100 mL in a volumetric flask.

7.3.7   Nitrate (NO"3-N) 1000 mg/L: Dissolve 0.6068 g sodium nitrate (NaNO3,
       CASRN 7631-99-4) in reagent water and dilute to 100 mL in a volumetric
       flask.

7.3.8   Nitrite (NO'2-N) 1000 mg/L: Dissolve 0.4926 g sodium nitrite (NaNO2,
       CASRN 7632-00-0) in reagent water and dilute to 100 mL in a volumetric
       flask.

7.3.9   Phosphate (PO43'-P) 1000 mg/L:  Dissolve 0.4394  g potassium
       dihydrogenphosphate (KH2PO4, CASRN 7778-77-0) in reagent water and
       dilute to 100 mL in a volumetric flask.

7.3.10 Sulfate (SO42') 1000 mg/L:  Dissolve 0.1814 g potassium sulfate (K2SO4,
       CASRN 7778-80-5) in reagent water and dilute to 100 mL in a volumetric
       flask.

NOTE:    Stability of standards:  Stock standards (7.3) for most anions are stable
           for at least 6 months when stored at 4°C. Except for the chlorite
           standard which is only stable for two weeks when stored protected from
           light at 4°C, and nitrite and phosphate which are only stable for 1 month
           when stored at 4°C. Dilute working standards should be prepared
           monthly, except those that contain chlorite, or nitrite and phosphate
           which should be prepared fresh daily.
                             300.1-11

-------
      7.4  Ethylenediamine (EDA) preservation solution, 100 mg/mL: Dilute 2.8 mL of
           ethylenediamine (99%) (CASRN 107-15-3) to 25 mL with reagent water. Prepare
           fresh monthly.

      7.5  Surrogate Solution: 0.50 mg/mL dichloroacetate (DCA) prepared by dissolving
           0.065 g dichloroacetic acid, potassium salt (C12CHCO2K, CASRN 19559-59-2) in
           reagent water and dilute to 100 mL in a volumetric flask.

           7.5.1   Dichloroacetate is potentially present in treated drinking waters as the
                   acetate of the organic disinfection by product, dichloroacetic acid (DCAA).
                   Typical concentrations of DCAA rarely exceed 50 ug/L, which, for this
                   worst case example, would represent only a five percent increase in the
                   observed response over the fortified concentration of 1.00 mg/L.
                   Consequently, the criteria for acceptable recovery (90% to 115%) for the
                   surrogate is weighted to 115% to allow for this potential background.

           7.5.2   Prepare this solution fresh every 3 months or sooner if signs of degradation
                   are present.

8.    SAMPLE COLLECTION. PRESERVATION AND STORAGE

      8.1   Samples should be collected in plastic or glass bottles.  All bottles must be
           thoroughly cleaned and rinsed with reagent water. Volume collected should be
           sufficient to insure a representative sample, allow for replicate analysis, if required,
           and minimize waste disposal.

      8.2   Special sampling requirements and precautions for chlorite.

           8.2.1    Sample bottles used for chlorite analysis must be opaque to protect the
                   sample from light.

           8.2.2   When preparing the LFM, be aware that chlorite is an oxidant and may react
                   with the natural organic matter in an untreated drinking water matrix as a
                   result of oxidative demand. If untreated water is collected for chlorite
                   analysis, and subsequently used for the LFM, EDA preservation will not
                   control this demand and reduced chlorite recoveries may be observed.

      8.3   Sample preservation and holding times for the anions that can be determined by this
           method are as follows:
                                       300.1-12

-------
      PART A :  Common Anions
      Analyte
      Bromide
      Chloride
      Fluoride
      Nitrate-N
      Nitrite-N
      ortho-Phosphate-P
      Sulfate
Preservation
None required
None required
None required
Cool to 4°C
Cool to 4°C
Cool to 4°C
Cool to 4°C
      PART B :  Inorganic Disinfection By-products
      Analyte                      Preservation
      Bromate                     50 mg/L EDA
      Bromide                     None required
      Chlorate                     50 mg/L EDA
Holding Time
28 days
28 days
28 days
48 hours
48 hours
48 hours
28 days
                          Holding Time
                          28 days
                          28 days
                          28 days
      Chlorite
50 mg/L EDA, Cool to 4 ° C  14 days
8.4   When collecting a sample from a treatment plant employing chlorine dioxide, the
      sample must be sparged with an inert gas (helium, argon, nitrogen) prior to addition
      of the EDA preservative at time of sample collection:

8.5   All four anions, in Part B, can be analyzed in a sample matrix which has been
      preserved with EDA.  Add a sufficient volume of the EDA preservation solution
      (Sect. 7.4) such that the final concentration is 50 mg/L in the sample. This would be
      equivalent to adding 0.5 mL of the EDA preservation solution to 1 L of sample.

8.6   EDA is primarily used as a preservative for chlorite. Chlorite is susceptible to
      degradation both through catalytic reactions with dissolved iron salts and reactivity
      towards free chlorine which exists as hypochlorous acid/hypochlorite ion in most
      drinking water as a residual disinfectant. EDA serves a dual purpose as a
      preservative for chlorite by chelating iron as well as any other catalytically destructive
      metal cations and removing hypochlorous acid/hypochlorite ion by forming an
      organochloroamine. EDA preservation of chlorite also preserves the integrity of
      chlorate which can increase in unpreserved samples as a result of chlorite
      degradation. EDA also preserves the integrity of bromate concentrations by binding
      with hypobromous acid/hypobromite which is an intermediate formed as by-product
      of the reaction of either ozone or hypochlorous acid/hypochlorite with bromide ion.
      If hypobromous acid/hypobromite is not removed from the matrix further reactions
      may form bromate ion.
                                  300.1-13

-------
      8.7   Degradation of ortho-phosphate has been observed in samples held at room
           temperature for over 16 hrs (see table 3 A). Therefore, samples to be analyzed for
           ortho-phosphate must not be held at room temperature for more than 12 cumulative
           hours.

9.    QUALITY CONTROL

      9.1   Each laboratory using this method is required to operate a formal quality control
           (QC) program. The requirements of this program consist of an initial demonstration
           of laboratory performance, and subsequent analysis in each analysis batch (Sect. 3.1)
           of a Laboratory Reagent Blank, Laboratory Fortified Blank, Instrument Performance
           Check Standard, calibration check standards, Laboratory Fortified Sample Matrices
           (LFM) and either Field, Laboratory or LFM duplicate sample analyses. This section
           details the specific requirements for each of these QC parameters. The laboratory is
           required to maintain performance records that define the quality of the data that are
           generated.

      9.2   INITIAL DEMONSTRATION OF PERFORMANCE

           9.2.1    The initial demonstration of performance is used to characterize instrument
                   performance (determination of accuracy through the analysis of the QCS)
                   and laboratory performance (determination of MDLs) prior to performing
                   analyses by this method.

           9.2.2   Quality Control Sample (QCS) — When beginning the use of this method, on
                   a quarterly basis or as required to meet data-quality  needs, verify the
                   calibration standards and acceptable instrument performance with the
                   preparation and analyses of a QCS. If the determined concentrations are not
                   within ± 15% of the stated values, performance of the determinative  step of
                   the method is unacceptable. The source of the problem must be identified
                   and corrected before either proceeding with the initial determination  of
                   MDLs or continuing with on-going analyses.

           9.2.3    Method Detection Limit (MDL) — MDLs must be established for all
                  • analytes, using reagent water (blank) fortified at a concentration of three to
                   five times the estimated instrument detection limit.(6) To determine MDL
                   values, take seven replicate aliquots of the fortified reagent water and
                   process through the entire analytical method over at least three separate
                   days.  Perform all calculations defined in the method and report the
                   concentration values in the appropriate units. Calculate the MDL as follows:
                                        300.1-14

-------
                               MDL = (t) x (S)

             where,    t =   Student's t value for a 99% confidence level and a standard
                            deviation estimate with n-1 degrees of freedom [t = 3.14 for
                            seven replicates].
                       S =  standard deviation of the replicate analyses.

             9.2.3.1    MDLs should be determined every 6 months, when a new
                       operator begins work or whenever there is a significant change in
                       the background, or instrument response.

9.3   ASSESSING LABORATORY PERFORMANCE

      9.3.1   Laboratory Reagent Blank (LRB) — The laboratory must analyze at least one
             LRB with each analysis batch (defined Sect 3.1). Data produced are used to
             assess contamination from the laboratory environment. Values that exceed
             the MDL indicate laboratory or reagent contamination should be suspected
             and corrective actions must be taken before continuing the analysis.

             9.3.1.1    If conducting analysis for the Part B anions, EDA must be added
                       to the LRB at 50 mg/L. By including EDA in the LRB, any bias
                       as a consequence of the EDA which may be observed in the field
                       samples, particularly in terms of background contamination, will
                       be identified.

      9.3.2   Laboratory Fortified Blank (LFB) - The LFB should be prepared at
             concentrations similar to those expected in the field samples and ideally at the
             same concentration used to prepare the LFM. Calculate accuracy as percent
             recovery (Sect. 9.4.1.3).  If the recovery of any analyte falls outside the
             required concentration dependant control limits (Sect! 9.3.2.2), that analyte
             is judged out of control, and the source of the problem should be identified
             and resolved before continuing analyses.

             9.3.2.1    If conducting analysis for the Part B anions, EDA must be added
                      to the LFB at 50 mg/L. The addition of EDA to all reagent water
                      prepared calibration and quality control samples is required not as
                      a preservative but rather as a means to normalize any bias
                      attributed by the presence of EDA in the field samples.

             9.3.2.2    Control Limits for the LRB

                 Concentration range                   Percent Recovery Limits
                  MRL to lOxMRL                    75 - 125 %
             lOxMRL to highest calibration level          85 - 115 %
                                  300.1-15

-------
            9.3 .2.2. 1    These control limits only apply if the MRL is established
                       within a factor of 10 times the MDL.  Otherwise, the limits
                       are set at 85% to 1 1 5%.

        9.3.2.3   The laboratory must use the LRB to assess laboratory perfor-
                 mance against the required control limits listed in 9.3.2.2. When
                 sufficient internal performance data become available (usually a
                 minimum of 20-30 analyses), optional control limits can be
                 developed from the percent mean recovery (x) and the standard
                 deviation (S) of the mean recovery. These data can be used to
                 establish the upper and lower control limits as follows:

                       UPPER CONTROL LIMIT = x + 3S
                       LOWER CONTROL LIMIT = x - 3S

                 The optional control limits must be equal to or better than those
                 listed in 9.3.2.2. After each five to ten new recovery measure-
                 ments, new control limits can be calculated using only the most
                 recent 20-30 data points. Also, the standard deviation (S) data
                 should be used to establish an on-going precision statement for the
                 level of concentrations monitored. These data must be kept on file
                 and be available for review.

9.3.3    Instrument Performance Check Solution (IPC) — The Initial Calibration
        Check Standard is to be evaluated as the instrument performance check
        solution in order to confirm proper instrument performance. Proper
        chromatographic performance must be demonstrated by calculating the Peak
        Gaussian Factor (PGF), which is a means to measure peak symmetry and
        monitoring retention time drift in the surrogate peak over time.  Critically
        evaluate the surrogate peak in the initial calibration check standard, and
        calculate the PGF as follows,

                           1.83  x W(l/2)
                    PGF = ______________________
                    i VJJL   ——™™™^™~™™™»»™™™™«««.»«™—
         where:  W( 1 /2) is the peak width at half height
                      W(l/10) is the peak width at tenth height

       9.3.3. 1    The PGF must fall between 0.80 and 1. 15 in order to demonstrate
                 proper instrument performance.
                            300.1-16

-------
             9.3.3.2   The retention time for the surrogate in the IPC must be closely
                       monitored on each day of analysis and throughout the lifetime of
                       the analytical column.  Small variations in retention time can be
                       anticipated when a new solution of eluent is prepared but if shifts
                       of more than 2% are observed in the surrogate retention time,
                       some type of instrument problem is present. Potential problems
                       include improperly prepared eluent,  erroneous method parameters
                       programmed such as flow rate or some other system problem. The
                       chromatographic profile (elution order) of the target anions
                       following an ion chromatographic analysis should closely replicate
                       the profile displayed in the test chromatogram that was shipped
                       when the column was purchased. As a column ages, it is normal to
                       see a gradual shift and shortening of retention times, but if after
                       several years of use, extensive use over less than a year, or use
                       with harsh samples, this retention time has noticeably shifted to
                       any less  than 80% of the original recorded value, the column may
                       require cleaning or replacement. Particularly if resolution
                       problems are beginning to become common between previously
                       resolved peaks.  A laboratory must retain a historic record of
                       retention times for the surrogate and all the target anions to
                       provide  evidence of an analytical columns vitality.

9.4   ASSESSING ANALYTE RECOVERY AND DATA QUALITY

      9.4.1   Laboratory Fortified Sample Matrix (LFM) — The laboratory must add a
             known amount of analyte to a minimum of 10% of the field samples within
             an analysis batch.  The LFM sample must be prepared from a sample matrix
             which has been analyzed prior to fortification. The analyte concentration
             must be high enough to be detected above the original sample and should
             adhere to the requirement of 9.4.1.2. It is recommended that the solutions
             used to fortify the LFM be prepared from the same stocks used to prepare
             the calibration standards and not from external source stocks. This will
             remove the bias contributed by an externally prepared stock and focus on any
             potential bias introduced by the field sample matrix.

             9.4.1.1    If the fortified concentration is less than the observed background
                       concentration of the unfortified matrix, the recovery should  not be
                       calculated. This is due to the difficulty in calculating accurate
                       recoveries of the fortified concentration when the native sample
                       concentration is so high.

             9.4.1.2    The LFM should be prepared at concentrations no greater than
                       five times the highest concentration observed in any field sample.
                       If no analyte is observed in any field sample, the LFM must be
                                  300.1-17

-------
                  fortified no greater than five times the lowest calibration level
                  which as outlined in 12.2 is the minimum reported level (MRL).
                  For example, if bromate is not detected in any field samples above
                  the lowest calibrations standard concentration of 5.00 ug/L, the
                  highest LFM fortified concentration allowed is 25.0 ug/L.

        9.4. 1.3    Calculate the percent recovery for each analyte, corrected for
                  concentrations measured in the unfortified sample.  Percent
                  recovery should be calculated using the following equation:

                               C8-C
                          R=  --------
                 where,       R =   percent recovery.
                              Cs =   fortified sample concentration
                              C =   sample background concentration
                              s  =    concentration equivalent of analyte added to
                                     sample.

        9.4. 1 .4   Until sufficient data becomes available (usually a minimum of 20
                 to 30 analysis), assess laboratory performance against recovery
                 limits of 75 to 125%. When sufficient internal performance data
                 becomes available develop control limits from percent mean
                 recovery and the standard deviation of the mean recovery. The
                 optional control limits must be equal to or better than the required
                 control limits of 75-125%.

        9.4. 1 .5   If the recovery of any analyte falls outside the designated LFM
                 recovery range and the laboratory performance for that analyte is
                 shown to be in control (Sect. 9.3),  the recovery problem
                 encountered with the LFM is judged to be either matrix or
                 solution related, not system related.

9.4.2    SURROGATE RECOVERY --  Calculate the surrogate recovery from all
        analyses using the following formula
                               SFC

       where,    R = percent recovery.
                 SRC = Surrogate Recovered Concentration
                 SFC = Surrogate Fortified Concentration
                            300.1-18

-------
       9.4.2.1    Surrogate recoveries must fall between 90-115% for proper
                 instrument performance and analyst technique to be verified. The
                 recovery of the surrogate is slightly bias to 115% to allow for the
                 potential contribution of trace levels of dichloroacetate as the
                 halogenated organic disinfection by-product (DBF) dichloroacetic
                 acid (DCAA) Background levels of this organic DBF are rarely
                 observed above 50 ug/L (0.05 mg/L) which constitutes only 5%
                 of the 1.00 mg/L recommended fortified concentration.

       9.4.2.2    If the surrogate recovery falls outside the 90-115% recovery
                 window, a analysis error is evident and sample reanalysis is
                 required'. Poor recoveries could be the result  of imprecise sample
                 injection or analyst fortification errors.

9.4.3   FIELD OR LABORATORY DUPLICATES » The laboratory must analyze
       either a field or a laboratory duplicate for a minimum of 10% of the collected
       field samples or at least one with every analysis batch, whichever is greater.
       The sample matrix selected for this duplicate analysis must contain
       measurable concentrations of the target anions in order to  establish the
       precision of the analysis set and insure the quality of the data. If none of the
       samples within an analysis batch have measurable concentrations, the LFM
       should be employed as a laboratory duplicate.

       9.4.3.1     Calculate the relative percent difference (RPD)  of the initial
                  quantitated concentration (Ic) and duplicate quantitated
                  concentration (Dc) using the following formula,

                              (IC-DC)
                      RPD =	X 100
                            (Pc + DJ/2)

        9.4.3.2    Duplicate analysis acceptance criteria

                  Concentration ranee                     %D)ff Limits
                   MRL to lOxMRL                      ± 20 %
                  lOxMRL to highest calibration level       ± 10 %

        9.4.3.3    If the %Diff fails to meet these criteria,  the samples must be
                  reanalyzed.

 9.4.4   Where reference materials are available, they should be  analyzed to provide
        additional performance data. The analysis of reference  samples is a valuable
        tool for demonstrating the ability to perform the method acceptably.
                              300.1-19

-------
            9.4.5   In recognition of the rapid advances occurring in chromatography, the
                   analyst is permitted certain options, such as the use of different columns,
                   injection volumes, and/or eluents, to improve the separations or lower the
                   cost of measurements. Each time such modifications to the method are
                   made, the analyst is required to repeat the procedure in Sect. 9.2 and adhere
                   to the condition of baseline stability found in Sect.  1.2.1.

            9.4.6   It is recommended that the laboratory adopt additional quality assurance
                   practices for use with this method. The specific practices that are most
                   productive depend upon the needs of the laboratory and the nature of the
                   samples.  Whenever possible, the laboratory should perform analysis of
                   quality control check samples and participate in relevant performance
                   evaluation sample studies.

10.   CALIBRATION AND STANDARDIZATION

      10.1  Establish ion chromatographic operating parameters equivalent to those indicated in
           Tables 1A or IB if employing a 2 mm column, Table 1C if employing a 4 mm
           column.

      10.2  Estimate the Linear Calibration Range (LCR) ~ The LCR should cover the expected
           concentration range of the field samples and should not extend over more than 2
           orders of magnitude in concentration (For example, if quantitating nitrate in the
           expected range of 1.0 mg/L to 10 mg/L, 2 orders of magnitude would permit the
           minimum and maximum  calibration standards of 0.20 mg/L and 20 mg/L,
           respectively.) The restriction of 2 orders of magnitude is prescribed since beyond this
           it is difficult to maintain linearity throughout the entire calibration range.

           10.2.1  If quantification  is desired over a larger range, then two separate calibration
                   curves should be prepared.

           10.2.2 For an individual calibration  curve, a minimum of three calibration standards
                   are required for a curve that  extends over a single order of magnitude and a
                  minimum of five calibration standards are required if the curve covers two
                  orders of magnitude. (For example, using the nitrate example cited above in
                  section 10.2,  but in this case limit the curve to extend only from 1.0 mg/L to
                   10 mg/L or a single order of magnitude. A third standard is required
                  somewhere in the middle of the range. For the calibration range of 0.20 mg/L
                  to 20 mg/L, over two orders of magnitude, five calibrations standards should
                  be employed, one each at the lower and upper concentration ranges and the
                  other three proportionally divided throughout the middle of the curve.)
                                        300.1-20

-------
10.3  Prepare the calibration standards by carefully adding measured volumes of one or
      more stock standards (7.3) to a volumetric flask and diluting to volume with reagent
      water.

      10.3.1  For the Part B anions, EDA must be added to the calibration standards at 50
             mg/L. The addition of EDA to all reagent water prepared calibration and
             quality control samples is required not as a preservative but rather as a means
             to normalize any bias attributed by the presence of EDA in the field samples.

      10.3.2  Prepare a 10.0 mL aliquot of surrogate fortified calibration solution which
             can be held for direct manual injection or used to fill an autosampler vial.
             Add 20 uL of the surrogate solution (7.5)  to a 20 mL disposable plastic
             micro beaker. Using a 10.0 mL disposable pipet, place exactly 10.0 mL of
             calibration standard into the micro beaker and mix. The calibration standard
             is now ready for analysis.  The same surrogate solution that has been
             employed for the standards should also be  used in the section 11.3.2 for the
             field samples.

10.4  Using a 2 mm column, inject 10 uL (Part A) or 50  uL (Part B) of each calibration
      standard. Using a 4 mm column, inject 50 uL (Part A) or 200 uL (Part B) of each
      calibration standard. Tabulate peak area responses against the concentration.  The
      results are used to prepare calibration curves using a linear least squares fit for each
      analyte. Acceptable calibration curves are confirmed after reviewing the curves for
      linearity and passing the  criteria for the initial calibration check standard in section
      10.5.1. Alternately, if the ratio of response to concentration (response factor) is
      constant over the LCR (indicated by < 15%  relative standard deviation (RSD),
      linearity through the origin can be assumed and the average ratio  or calibration factor
      can be used in place of a calibration curve,

      10.4.1  Peak areas are strongly recommended since they have been found to be more
             consistent, in terms of quantitation,  than peak heights.  Peak height can tend
             to be suppressed as a result of high levels of common anions in a given
             matrix which can compete for exchange sites. Using peak areas, it is the
             analyst responsibility to review all chromatograms to insure accurate baseline
             integration of target analyte peaks since poorly drawn baselines will more
             significantly influence peak areas than peak heights.

10.5  Once the calibration curves  have been established they must be verified prior to
      conducting any sample analysis using an initial calibration check standard (3.2.2).
      This verification must be performed on each analysis day or whenever fresh eluent
      has been prepared. A continuing calibration check standard (3.2.3) must be analyzed
      after every tenth sample  and at the end of the analysis set as an end calibration check
      standard (3.2.4). The response for the initial, continuing and end calibration check
      must satisfy the criteria listed in  10.5.1. If during the analysis set, the response differs
                                   300.1-21

-------
            by more than the calibration verification criteria shown in 10.5.1., or the retention
            times shift more than ± 5% from the expected values for any analyte, the test must be
            repeated, using fresh calibration standards. If the results are still outside these
            criteria, sample analysis must be discontinued, the cause determined and/or in the
            case of drift, the instrument recalibrated. All samples following the last acceptable
            calibration check standard must be reanalyzed.

            10.5.1  Control limits for calibration verification

                       Concentration range                    Percent Recovery Limits
                         MRL to lOxMRL                     75 - 125 %
                   1 OxMRL to highest calibration level          85 - 115 %

                    10.5.1.1 These control limits only apply if the MRL is established within a
                            factor of 10 times the MDL.  Otherwise, the limits are set at 85%
                            to 115%.

            10.5.2  CALIBRATION VERIFICATION REQUIREMENT FOR PART B

                   As a mandatory requirement of calibration verification, the laboratory MUST
                   verify calibration using the lowest calibration standard as the initial
                   calibration check standard.

            10.5.3  After satisfying the requirement of 10.5.2, the levels selected for the  other
                   calibration check standards should be varied between a middle calibration
                   level and the highest calibration level.

11.   PROCEDURE

      11.1  Tables 1A and IB summarize the recommended operating conditions for the  ion
           chromatograph.  Included in these tables are estimated retention times that can be
           achieved by this method.  Other columns,  chromatographic conditions, or detectors
           may be used if the requirements of Sect. 9.2 are met.

      11.2  Check system calibration  daily and, if required, recalibrate as described in Sect. 10.

      11.3  Sample Preparation

           11.3.1  For refrigerated or samples arriving to the laboratory cold, ensure the
                  samples have come to room temperature prior to conducting sample  analysis
                  by allowing the samples to warm on the bench for at least 1 hour.

           11.3.2  Prepare a 10.0 mL aliquot of surrogate fortified sample which can be held for
                  direct manual injection or used to fill an autosampler vial.  Add 20 uL of the


                                        300.1-22

-------
              surrogate solution (7.5) to a 20 mL disposable plastic micro beaker. Using a
              10.0 mL disposable pipet, place exactly 10.0 mL of sample into the micro
              beaker and mix. Sample is now ready for analysis.

              11.3.2.1  The less than 1% dilution error introduced by the addition of the
                       surrogate is considered insignificant.

11.4  Using a Luer lock, plastic 10 mL syringe, withdraw the sample from the micro
      beaker and attach a 0.45 \im paniculate filter (demonstrated to be free of ionic
      contaminants)  directly to the syringe.  Filter the sample into an autosampler vial (If
      vial is not designed to automatically filter) or manually load the injection loop
      injecting a fixed amount of well mixed sample. If using a manually loaded injection
      loop, flush the loop thoroughly between sample analysis using sufficient volumes of
      each new sample matrix.

11.5  Using a 2 mm  column, inject 10  uL (Part A) or 50 uL (Part B) of each sample.
      Using a 4 mm  column, inject 40  uL (Part A) or 200 uL (Part B) of each sample.
      Tabulate peak  area responses against the concentration. During this procedure,
      retention times must be recorded. Use the same size loop for standards and samples.
      Record the resulting peak size in area units.  An automated constant volume injection
      system may also be used.

11.6  The width of the retention time window used to make identifications should be based
      upon measurements of actual retention time variations of standards over the course
      of a day. Three times the standard deviation of a retention time can be used to
      calculate a suggested window size for each analyte. However, the experience of the
      analyst should  weigh  heavily in the interpretation of chromatograms.

11.7  If the response of a sample analyte exceeds the calibration range, the sample may be
      diluted with an appropriate amount of reagent water and reanalyzed. If this is not
      possible then three new calibration concentrations must be employed to create a
      separate high concentration curve, one standard near  the estimated concentration
      and the other two bracketing around an interval equivalent to ± 25% the estimated
      concentration.  The latter procedure involves significantly more time than a simple
      sample dilution therefore, it is advisable to collect sufficient sample to  allow for
      sample dilution or sample reanalysis, if required.

11.8  Shifts in retention time are inversely proportional to concentration.  Nitrate,
      phosphate and  sulfate will exhibit the greatest degree of change, although all anions
      can be affected. In some cases this peak migration may produce poor  resolution or
      make peak identification difficult.

11.9  Should more complete resolution be needed between any two coeluting peaks, the
      eluent (7.2) can be diluted. This will  spread out the run, however, and will cause late
                                   300.1-23

-------
           eluting anions to be retained even longer. The analyst must determine to what extent
           the eluent is diluted.  This dilution is not be considered a deviation from the method.
           If an eluent dilution is performed, section 9.2 must be repeated.

           11.9.1   Eluent dilution will reduce the overall response of an anion due to
                    chromatographic band broadening which will be evident by shortened and
                    broadened peaks.  This will adversely effect the MDLs for each analyte.

12.   DATA ANALYSIS AND CALCULATIONS

      12.1  Prepare a calibration curve for each analyte by plotting instrument response, as peak
           area, against standard concentration. Compute sample concentration by comparing
           sample response with the standard curve.  If a sample has been diluted, multiply the
           response by the appropriate dilution factor.

      12.2  Report ONLY those values  that fall between the lowest and the highest calibration
           standards. Samples with target analyte responses exceeding the highest standard
           should be diluted and reanalyzed. Samples with target  analytes identified but
           quantitated below the concentration established by the lowest calibration standard
           should be reported as below the minimum reporting limit (MRL).

      12.3  Report results for Part A anions in mg/L and for Part B anions in ug/L.

      12.4  Report   NO2'  asN
                    NO3"  asN
                    HPO4= as P
                    Br" in mg/L when  reported with Part A
                    Br" in ug/L when reported with Part B

13.   METHODS PERFORMANCE

      13.1  Tables 1 A, IB, and 1C give the single laboratory (OW  OGWDW TSC-Cincinnati)
           retention times, standard conditions and MDL determined for each anion included in
           the method.  MDLs for the Part A anions were determined in reagent water on the 2
           mm column  (Table 1 A).  MDLs for the Part B anions were conducted not only in
           reagent water but also a  simulated high ionic strength water (HTW) on the 2 mm
           column (Table IB) and in reagent water on the 4 mm column (Table 1C). HTW is
           designed to simulate a high ionic strength field sample.  It was prepared from reagent
           water which was fortified with the common anions of chloride at 100 mg/L,
           carbonate at 100 mg/L, nitrate at 10.0 mg/L as nitrogen, phosphate at 10.0  mg/L as
           phosphorous, and sulfate at  100 mg/L.
                                        300.1-24

-------
      13.2 Tables 2 A and 2B give the single laboratory (OW OGWDW TSC-Cincinnati)
           standard deviation for each anion included in the method in a variety of waters for the
           standard conditions identified in Table 1A and IB, respectively.

     13.3  Tables 3 A and 3B shown stability data for the Part A and B anions, respectively.
           Each data point in these tables represent the mean percent recovery following
           triplicate analysis. These data were used to formulate the holding times shown in
           Sect. 8.3.

14.  POLLUTION PREVENTION

     14.1  Pollution prevention encompasses any technique that reduces or eliminates the
           quantity or toxicity of waste at the point of generation.  Numerous opportunities for
           pollution prevention exist in laboratory operation. The EPA has established a
           preferred hierarchy of environmental management techniques that places pollution
           prevention as the management option of first choice. Whenever feasible, laboratory
           personnel should use pollution prevention techniques to address their waste
           generation.  When wastes cannot be feasibly reduced at the source, the Agency
           recommends recycling as the next best option.

     14.2  Quantity of the chemicals purchased should be based on expected usage during its
           shelf life and disposal cost of unused material. Actual reagent preparation volumes
           should reflect anticipated usage and reagent stability.

     14.3  For information about pollution prevention that  may be applicable to laboratories
           and research institutions, consult "Less is Better: Laboratory Chemical Management
           for Waste Reduction," available from the American Chemical Society's Department
           of Government Regulations and Science Policy, 1155 16th Street N.W., Washington
           D.C. 20036,
           (202) 872-4477.

15.  WASTE MANAGEMENT

     15.1  The Environmental Protection Agency requires that laboratory waste management
           practices be conducted consistent with all applicable rules and regulations.  Excess
           reagents, samples and method process wastes should be characterized and disposed
           of in an acceptable manner. The Agency urges laboratories to protect the air, water,
           and land by minimizing and controlling all releases from  hoods  and bench operations,
           complying with the letter and spirit of any waste discharge permit and regulations,
           and by complying with all solid and hazardous waste regulations, particularly the
           hazardous  waste identification rules and land disposal restrictions. For further
           information on waste management consult the "Waste Management Manual for
           Laboratory Personnel," available from the American Chemical Society at the address
           listed in Sect. 14.3.
                                        300.1-25

-------
16.   REFERENCES

     1.     "Determination of Inorganic Disinfection By-Products by Ion Chromatography", J.
           Pfaff, C. Brockhoff.  J. Am. Waterworks Assoc., Vol 82, No. 4, pg 192.

     2.     Standard Methods for the Examination of Water and Wastewater, Method 411 OB,
           "Anions by Ion Chromatography", 18th Edition of Standard Methods (1992).

     3.     Dionex, System DX500 Operation and Maintenance Manual, Dionex Corp.,
           Sunnyvale, California 94086, 1996.

     4.     Method Detection Limit (MDL) as described in "Trace Analyses for Wastewater," J.
           Glaser, D. Foerst, G. McKee, S. Quave, W. Budde,   Environmental Science and
           Technology, Vol. 15, Number 12, page 1426, December, 1981.

     5.     American Society for Testing and Materials. Test Method for Anions in Water by
           Chemically-Suppressed Ion Chromatography D4327-91. Annual Book of Standards
           Vol 11.01  (1993).

     6.     Code of Federal Regulations 40, Ch. 1, Pt.  136, Appendix B.

     7.    Hautman, D.P. & Bolyard, M. Analysis of Oxyhalide Disinfection By-products and
          other Anions of Interest in Drinking Water by Ion Chromatography. Jour, of
          Chromatog., 602, (1992), 65-74.

    8.     Standard Methods for the Examination of Water and Wastewater, Method 4500-
          C1O 2,C " Amperometric Method I" (for the determination of Chlorine Dioxide), 19th
          Edition of Standard Methods (1995).
                                     300.1-26

-------
17.  TABLES. DIAGRAMS. FLOWCHARTS AND VALIDATION DATA
TABLE 1A. CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION
            LIMITS IN REAGENT WATER FOR THE COMMON ANIONS (PART A).

ANALYTE
Fluoride
Chloride
Nitrite-N
Surrogate: DCA
Bromide
Nitrate-N
ortho-Phosphate-P
Sulfate

PEAK # (1>
1
2
3
4
5
6
7
8

RETENTION TIME
(MIN.)
2.53
4.67
6.01
7.03
8.21
9.84
11.98
13.49
MDL
Fort Cone,
mg/L
0.020
0.020
0.010

0.040
0.010
0.040
0.040
1
DETERMINATION
Number
of
Replicates
7
7
7

7
7
7
7
DI
MDL
mg/L
0.009
0.004
0.001

0.014
0.008
0.019
0.019
Standard Conditions:

Ion Chromatograph:
Columns:
Detector:
Suppressor:
Eluent:
Eluent Flow:
Sample Loop:
Dionex DX500
Dionex AG9-HC / AS9-HC, 2 mm
Suppressed Conductivity Detector, Dionex CD20
ASRS-I, external source electrolytic mode, 100 mA current
9.0 mM Na2CO3
0.40 mL/min
10 uL
System Backpressure:  2800 psi
Background Conductivity:      22 uS

Recommended method total analysis time:  25 minutes

(1)  See Figure 1
                                    300.1-27

-------
TABLE IB.    CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION
               LIMITS IN BOTH REAGENT WATER AND HIGH IONIC STRENGTH
               WATER FOR THE INORGANIC DISINFECTION BY-PRODUCTS
               (PART B).


ANALYTE
Chlorite
Bromate
Surrogate:
DCA
Bromide
Chlorate


PEAK # (1)
1
2
4
5
6
.

RETENTION
TIME
(MIN.)
3.63
4.19
7.28
8.48
9.28
MDL DETERMINATION
Fort
Cone,
Hg/L
2.00
2.00

2.00
2.00
Number
of
Replicates
7
7

7
7
DI
MDL
0.89
1.44

1.44
, 1.31
Hr\v<2)
MDL
"g/L
0.45
1.28

2.51
0.78
Standard Conditions:

Ion Chromatograph:
Columns :
Detector:
Suppressor:
Eluent:
Eluent Flow:
Sample Loop:
Dionex DX500
Dionex AG9-HC / AS9-HC, 2 mm
Suppressed Conductivity Detector, Dionex CD20
ASRS-I, external source electrolytic mode, 100 mA current
9.0mMNa2CO3
0.40 mL/min
50  L
System Backpressure: '2800 psi
Background Conductivity:      22 uS

Recommended method total analysis time: 25 minutes

(1)  See Figure 2 and 3

(2)   HIW indicates High Ionic Strength Water which is a simulated drinking water prepared
     from reagent water and fortified with chloride at 100 mg/L, carbonate at 100 mg/L, nitrate
     at 10.0 mg/L as nitrogen, phosphate at 10.0 mg/L as phosphorous, and sulfate at 100 mg/L.
                                      300.1-28

-------
TABLE 1C.   CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION
              LIMITS IN REAGENT WATER FOR THE INORGANIC
              DISINFECTION BY-PRODUCTS USING AN ALTERNATE 4 mm AS9-
              HC COLUMN (PART B).

ANALYTE
Chlorite
Bromate
Surrogate:
DCA
Bromide
Chlorate

PEAK#
1
2
4
5
6

RETENTION
TIME
(MIN.)
4.43
5.10
8.82
10.11
10.94
MDL DETERMINATION
Fort
Cone,
Hg/L
2.00
2.00

2.00
2.00
Number
of
Replicates
7
7

7
7
DI
MDL
ug/L
1.44
1.32

0.98
2.55
Standard Conditions:

Ion Chromatograph:
Columns:
Detector:
Suppressor:
Eluent:
Eluent Flow:
Sample Loop:
Dionex DX500
Dionex AG9-HC / AS9-HC, 4 mm
Suppressed Conductivity Detector, Dionex CD20
ASRS-I, external source electrolytic mode, 300 mA current
1.25 mL/min
200 uL
System Backpressure:  1900 psi
Background Conductivity:      21 uS

Recommended method total analysis time: 25 minutes
                                    300.1-29

-------
  TABLE 2A.    SINGLE-OPERATOR PRECISION AND RECOVERY FOR THE COMMON
ANALYTE
Fluoride



Chloride



Nitrite-N



Bromide
•


Nitrate-N



Phosphate-P



Sulfate



Surrogate:



ANIONS (PART A).
UNFORT
MATRIX
MATRIX CONC, .
mg/L
RW
SW
GW
CDW
RW
SW
GW
CDW
RW
SW
GW
CDW
RW
SW
GW
CDW
RW
SW
GW
CDW
RW
SW
GW
CDW
RW
SW
GW
CDW
RW
SW
GW
CDW

-------
  TABLE 2B.   SINGLE-OPERATOR PRECISION AND RECOVERY FOR THE
                INORGANIC DISINFECTION BY-PRODUCTS (PART B).

Chlorite RW

fflW

SW

GW

C1W

CDW

O3W

Bromate RW

raw

SW

GW

C1W

CDW

O3W

UNFORT
CONC.
ug/L

-------
TABLE 2B. SINGLE-OPERATOR PRECISION AND RECOVERY FOR THE INORGANIC
           DISINFECTION BY-PRODUCTS (PART B) (contd.).
UNFORT FORT #
CONC. CONG OF MEAN MEAN
ANALYTE MATRIX ug/L ug/L REPLC ug/L %REC




























RW
fflW
Bromide RW

fflW

SW

GW

C1W

CDW

O3W

Chlorate RW

fflW
.
SW

GW

C1W

CDW

O3W

= Reagent Water
= High Ionic strength

-------
TABLE 2B.   SINGLE-OPERATOR PRECISION AND RECOVERY FOR THE INORGANIC
              DISINFECTION BY-PRODUCTS (PART B)(contd.).
ANALYTE
Surrogate: DCA
(see NOTE below)











MATRIX
RW
fflW

SW

GW

C1W

CDW

O3W

FORT
CONC
mg/L
5.00
5.00

5.00

5.00

5.00

5.00

5.00

#
OF
REPLC
9
9

9

9

9

9

9

MEAN
mg/L
5.11
4.98
5.00
4.96
4.95
4.99
5.12
5.13
5.15
5.13
5.01
5.04
4.99
5.11
MEAN
%REC
102
99.5
100
99.2
98.9
99.8
102
103
103
103
100
101
99.8
101
SD(n-l)
0.93
0.69
0.79
1.76
0.70
1.60
0.50
0.50
1.73
1.12
1.02
1.08
0.70
0.53
%RSD
0.91
0.69
0.79
1.78
0.7
1.61
0.49
0.49
1.68
1.09
1.02
1.07
0.7
0.52
 RW = Reagent Water
HTW = High Ionic strength Water
      [see note (2) in Table IB]
SW= Surface Water
                       GW = Groundwater
                       C1W = Chlorinated drinking water
                      CDW = Chlorine dioxide treated drinking water
                      O3W = Ozonated drinking water
NOTE:
The surrogate DCA was fortified at 5 mg/L but due to concerns about measuring trace
concentrations of bromide with such high concentration of the neighboring surrogate peak,
the recommended fortified concentration for the surrogate has been reduced to 1.00 mg/L.
                                          300.1-33

-------
TABLE 3A.    STABILITY STUDY RESULTS FOR THE COMMON ANIONS (PART A).
ANALYTE
Fluoride




Chloride



Nitrite-N



Bromide



Nitrate-N



Phosphate-P



Sulfate



Preservative Matrix
None RW

SW
GW
CDW
None RW
SW
GW
CDW
None RW
SW
GW
CDW
None RW
SW
GW
CDW
None RW
SW
GW
CDW
None RW
SW
GW
CDW
None RW
SW
GW
CDW
UNFORT
CONC.
mg/L

-------
TABLE 3B STABILITY STUDY RESULTS FOR THE INORGANIC
DISINFECTION BY-
PRODUCTS (PARTB).

ANALYTE
Chlorite






Chlorite






Bromate






Bromate







Preservative Matrix
None RW
raw
SW
GW
C1W
CDW
O3W
EDA RW
HIW
SW
GW
C1W
CDW
O3W
None RW
HIW
SW
GW
C1W
CDW
O3W
EDA RW
HIW
SW
GW
C1W
CDW
O3W
UNFORT
CONG,
Hg/L

-------
* TABLE 3B. STABILITY STUDY RESULTS FOR THE INORGANIC DISINFECTION
 BY-PRODUCTS (PART B)(contd.)

ANALYTE

Bromide







Bromide






Chlorate






Chlorate








Preservative Matrix

None RW
fflW
SW
GW

C1W
CDW
O3W
EDA RW
fflW
SW
GW
C1W
CDW
O3W
None RW
HIW
SW
GW
C1W
CDW
O3W
EDA RW
HIW
SW

GW
C1W
CDW
O3W
UNFORT
CONC.


-------
   o   o o o o o
  |CM o CM o CM CM o co

  too CM' co 10 co' co co' co'
  :!  co              CO
                            .

                            -ti
                             ™
   tO CD p O CM CO CO •sf

   CM' TJ-' CD" f~" co' co' T^ co
 rohr- CM co Tf m CD h- co

 $
Q.  '
IO
                          £
                                                          CO
           CN
            o
            o
                                                                            o
                                                                            p

                                                                            cvi
                                                                            o
                                                                            o
o

P  CO
CO  

o

(0

o




I
        •c
        (0
        Q.

        0)
        £Z
                    o


                    'E
                    (0
                    Q.
                     g



                    II
                    co p
                    o o

                    CB .55
                    E «
                                                                                    O ro
         §
         D)

        LL.

-------
               28
c-tc^.^Sz  -Iro-E
 CO COO) CO
CO T- 03 CM -M- CM
             Pl§
               ??I<1
co
        oo at
T- CM co -^ m to
                                        CD
                                                       0
                                                       0
                                                      0
                                                      0

                                                   _  cvi
                                                        o
                                                      - p

                                                    _   o
                                                      o
                                                      o
                                                      0
                                                      o
                                                      o
                       CO
                                                             

                                                           L-
                                                                    00

-------
CO CO CO CO CO CO
    - 00 CM ^ CM

-------
llsfe
         «i  1
         -l£ v to
  o
o  ooo  oj^f
CM  T- T- *-  -»TH
          a   S
         a) g ^2 0) Q-
QQ


_02
JD


1=

0)
0)
CO

-------

-------