United States
                  Environmental Protection
                  agency
 Atmospheric Research and Exposure
 Assessment Laboratory
 Research Triangle Park, NC 27711
                  Research and Development
 EPA/600/S3-90/009  May 1990
&EPA        Project Summary
                  Laboratory and Field
                  Evaluations of Methodology
                  for Measuring Emissions of
                  Chlorinated  Solvents from
                  Stationary Sources

                  Anna C. Carver, William G. DeWees, and Easter A. Coppedge
                   An evaluation of the use of EPA
                 Method 18 for sampling and analysis
                 of chlorinated  solvents was needed
                 to support possible future regulations
                 by the U. S. Environmental Protection
                 Agency. The solvents specifically
                 addressed are: carbon tetrachloride
                 (CCI4),  chloroform   (CHCI3),
                 perchloroethylene  (PERC), and
                 trichloroethylene (TCE). Laboratory
                 and field studies were performed to
                 evaluate sampling and  analytical
                 procedures  for measuring  these
                 solvents from  stationary sources.
                 Conclusions  and recommendations
                 are made regarding the application of
                 Method 18  to organic  solvent
                 sampling and analysis.
                   This  Project  Summary  was
                 developed by  EPA's  Atmospheric
                 Research and Exposure Assessment
                 Laboratory, Research  Triangle Park,
                 NC, to announce key  findings of the
                 research project that  is fully
                 documented  in  a separate report of
                 the  same title  (see Project Report
                 ordering information at back).


                 Introduction
                   Several chlorinated solvents are being
                 evaluated by  the U. S. Environmental
                 Protection Agency (EPA) for possible
                 future regulations for emissions from
                 stationary sources.  Entropy Environ-
                 mentalists,  Inc.  (Entropy)  was
commissioned by the Quality Assurance
Division (QAD)  of EPA's Atmospheric
Research and  Exposure Assessment
Laboratory (AREAL) to evaluate methods
for sampling and analysis of some of the
high priority solvents. Those  solvents
included in the  evaluation were carbon
tetrachloride (CCI4) chloroform  (CHCI3),
perchloroethylene  (PERC),  and
trichloroethylene (TCE). The work was
conducted in two phases: (1)  a  field
evaluation of sampling and analytical
techniques for  measuring emissions of
PERC, arid (2)  laboratory and field
evaluations of the collection of all four
solvents under  high  temperature/high
moisture conditions.
  The  use of  EPA Method  181  was
previously evaluated as a sampling and
analysis method for PERC emissions in a
laboratory study  sponsored by the EPA.2
Since  perchloroethylene is widely
employed as a solvent in the chemical
degreasing industry, a degreasing facility
was selected for field evaluation of the
candidate method. Gaseous  samples
collected in Tedlar bags were analyzed
directly using  a gas  chromatograph
equipped with a flame ionization detector
(GC/FID).  Emission samples were  also
adsorbed onto  activated charcoal in
sorbent tubes and likewise analyzed by
GC/FID after  desorption with an
appropriate solvent.
  Laboratory evaluation of EPA Method
18 techniques modified to include an in-

-------
line gas  conditioning system  for  high
temperature/high  moisture  sample
collection involved (1) the identification of
adsorption media that  will  effectively
collect organic  solvents  under these
conditions and (2) the effectiveness of the
gas conditioning  system  in permitting
sample collection in Tedlar bags. A field
evaluation of the sorbents chosen and the
gas conditioning system developed was
conducted at a gas-fired  boiler where the
exhaust gas stream was spiked with the
chlorinated solvents. Samples  collected
both  in  Tedlar  bags  and  on  the
adsorption  tubes  were analyzed using
GC/FID.

Experimental Procedures
  All samples were analyzed using  a
Hewlett-Packard  5890A series  gas
chromatograph with dual flame  ionization
detectors.  One  injection  port  was
equipped  for liquid injections  while the
other was equipped with a  manually
operated gas sampling  valve. Column
selection and operating procedures  were
based  on  those followed in  previous
evaluations.3  For all  analyses,  the
hydrogen  and air pressures  supplied to
the dual FID's were maintained  at 22 and
20 psi, respectively. These values  were
selected by optimizing the FID response
according to manufacturer's suggestions.
For  the  GC/FID analysis,  calibration
standards were  prepared,  in bags with
nitrogen or in the  appropriate  desorbing
solution, using  measured amounts of
each solvent. These were injected into
the GC to obtain a linear calibration curve
in the  range of  the  samples to  be
analyzed.
  Dual quadruplicate-train  sampling runs
were conducted at the inlet and outlet to
a carbon bed adsorber at the degreasing
facility according  to  EPA Method  18.
Gaseous  samples  were  collected
simultaneously  in  four Tedlar  bags and
four  charcoal adsorption tubes. Double-
seamed Tedlar bags were fabricated for
the collection of gaseous samples  as
specified,  and each  bag  was  leak-
checked before and after use in the field.
Adsorption tubes were  packed  with
charcoal and checked for migration  of
PERC from the primary to  the  backup
portion.  The  sampling  flow rate  for
collection of the Tedlar bag samples was
maintained  using adjustable  needle
valves at a rate of 55 - 60  cms/min, while
the charcoal tube samples  were collected
at a flow rate of 0.08 L/min. also  using
critical orifices. The gaseous Tedlar bag
samples were  injected  directly into  the
GC/FID  and quantified  according  to
Method 18. The adsorbed charcoal tube
samples  were desorbed  using  a fixed
volume of methylene  chloride to allow
GC/FID analysis  using liquid injection of
the resulting solutions.
  Prior to the second field evaluation,
seven   commercially  available
sorbents(Bio-Rad's Bio BeadsR SM-2 and
SM-4; Supelco's Amberlite  XAD-8«
CarbotrapR and Carboxen; Water's C18-
OctadecylR and  Union  Carbide's TRI-X-
100R) were evaluated for collecting  the
solvents of interest under simulated high
temperature/high moisture conditions.  All
sorbents  were pretreated according  to
manufacturer's instructions.  Two-gram
quantities of  each  were packed into
sampling tubes  and   subjected  to
retention  and breakthrough (time  at which
sorbent  is no longer  retaining  one  or
more of the solvents) tests. Sorbent tubes
were challenged  with a mixture  of CCI4,
CHCI3, PERC,  and TCE each at 100 ppm
from a Tedlar bag at an approximate flow
of 1.0 L/min, for a 1-hour period, using a
diaphragm pump. The gas  exiting  the
tubes was introduced directly  into  the
GC/FID.
  High temperature/high  moisture stack
gas containing the chlorinated  solvents
was  simulated by  drawing the  same
mixture  at the same flow for the same
time period  through  a heated impinger
containing 400 ml of water, This  gas was
then drawn through the candidate in-line
gas conditioning  system condenser  for
cooling   and  a primary  sorbent tube,
followed by a knockout jar for condensate
collection and  a second (backup) sorbent
tube.  The pressure  differential of  the
system  was monitored,  as was  the  gas
flow rate. The  resulting sample was
introduced directly into the GC/FID.
  The two  sorbents  with  acceptable
retention and  breakthrough   chara-
cteristics under the  simulated  high
temperature/high moisture  conditions,
were  examined for efficiency of solvent
removal for  analysis (desorption
efficiency). The desorption efficiency was
determined for three different desorbing
solutions: acetonitrile,  methanol, and
methylene chloride.  Multiple tubes  of
each  sorbent  were  charged with  1.0 11!
each  of the  chlorinated solvents and  set
aside for 1 h. Using  a separate tube for
each  sorbent/desorbing solution
combination,  they were  desorbed with 4
ml of the appropriate  solution which was
then  analyzed by GC/FID. The  percent
desorption efficiency for each  solvent
from  each sorbent was calculated based
on the known charge of that solvent.
   To apply Tedlar bag sampling  to a high
temperature/high  moisture source, an in-
line  gas conditioning  system was
designed.  The  system consisted of  a
condenser to  cool the  gas  stream,
followed by a knockout jar to collect the
moisture, and then the Tedlar bag. In  a
laboratory  check of this system design,
condensate collected in the knockout jars
was  analyzed.  High  temperature/high
moisture  stack  gas  containing the
solvents was simulated  as previously
described and  drawn  through the
sampling system  at approximately  1.0
L/min for 1  h. Aliquots of the condensate
collected were  passed through a C18
cartridge,  eluted  with methanol, and
injected into the GC/FID, or placed into 4
ml of  methylene  chloride and  injected
into the GC/FID.
  An  emission  spiking  system  was
developed to  introduce known quantities
of the solvents into a stack with relatively
clean  emissions.  A  mixture  of  the
solvents equivalent  to the desired stack
gas concentrations was passed through a
pressurized atomizer to create a fine mist
that  would be  vaporized  in the  high
temperature  emissions.  A SpeedaireR
paint  sprayer was modified to include a
deVilbus nebulizing nozzle. The  system
pressure was  used to regulate the flow. A
Teflon flow meter was placed in-line for
fine adjustment of the flow rate.
  A preliminary test run was conducted
to determine any need for modification  in
the  sampling  system.  Then field
evaluation testing  using  the  refined
protocols  was conducted according  to
EPA Method  18 with modification of the
sampling  systems to include in-line
conditioning for the high temperature/high
moisture  gas  stream.  Each  of five
sampling runs was conducted over a 1-h
time  period.  The  spiking  system was
started 5 minutes before sampling began,
and  sampling continued 5 minutes after
all  of the solvent mixture had been
introduced into  the  stack. The sampling
probe, which was constructed  of Teflon
tubing, was inserted into  a port with the
inlet  at  the  center of  the  stack
approximately 18  feet downstream of the
point  of injection  of  the  solvents.  A
diaphragm pump was used to  pull the
sample gas  into a  heated sampling
manifold constructed  of  stainless steel
piping. Twenty  sampling  lines  were
attached  to the  manifold,  allowing
generation of  up  to  20 simultaneous
samples during a  single  sampling run
(see Figure 1 for a  detailed schematic of
an eight-train setup). Eight of the sample
lines  were equipped for Tedlar  bag
sampling, while the other  twelve were
equipped  for sorbent  tube sampling. A
sampling rate less than that specified  by
Method 18 was used for 12 of the  16 bag

-------
samples collected. To validate the use of
this reduced flow rate, one set of four bag
samples was collected at the higher rate.
The actual  flow rates  for the  small bag
samples were  maintained  (using critical
orifices) between 0.038 and 0.040 L/min,
while the higher flow rates  were 0.440 to
0.470  L/min. The flow rates through the
primary  and  backup sorbent tubes
ranged between  0.230 and 0.240 L/min,
also maintained by critical  orifices. The
resulting  Tedlar bag samples  were
analyzed within 4 hours of  sampling, and
the sorbent tube samples were analyzed
within  5 days of sampling.

Results and Discussion
  The  optimum operating  conditions
identified for analysis  of PERC samples
collected in Tedlar bags during the first
field  evaluation  utilized  a  packed
chromatography column containing  20%
SP-2100/0.1%  Carbowax  1500 on
100/120  mesh Supelcoport, operated
isothermally at 120°C. PERC  solutions
obtained from the desorption  of the
charcoal tubes using methylene chloride
were analyzed using an OV-101  column
and  the  following  temperature
programming:  70 °C (3 minutes)-ramp of
30°C/min (1 minute) - 100°C (3 minutes).
For  the  PERC   bag  samples,   the
estimated limit of detection was 1.6  ppm,
and the estimated  quantifiable  limit was
4.8 ppm. No  upper detection  limit was
apparent below the level at  which  room
temperature nitrogen  is  saturated  with
perchloroethylene. For liquid  injections of
4.0 ill, dissolved PERC in amounts as low
as 0.003 pi per 4  ml  desorbing solution
resulted  in  discernable  detector
response, while concentrations above
160  ul per 4  ml of desorbing  solution
saturated  the  detector.   Migration  of
PERC  from  the  primary  to the backup
portion of the  charcoal tubes  was found
to be insignificant at charge levels from
1.3 to 36.0 nL
  All  bag samples  from  the  first
evaluation were initially analyzed on site.
The  average  coefficient  of variation
observed for the bag samples from the
controlled emissions of PERC was 9.8%.
A  pooled  standard deviation of  14 ppm
was  calculated for charcoal tube  samples
collected from the controlled  emissions.
The  average  coefficient of variation  for
these samples  was 4.6%.
  For the second field test, optimization
of instrument  operating  conditions  for
                                                                                            3/8" Teflon Tubing
       z
                •3/8" Teflon Tubing
Figure 1.    Detailed schematic of multiple bagisorbent tube sampling train system.

-------
analysis of bag and desorbed sorbent
tube samples resulted in analysis of both
types using a packed chromatography
column  of 1%  SP-1000  on  60/80
Carbopack-B  Supelcoport, operated
isolhermally at 200°C. The carrier gas
How was maintained at 20 mL/min  for
both the liquid and the gaseous sample
injections.
  Of the seven sorbents  examined  for
chlorinated  solvent collection under high
temperature/high  moisture conditions,
two were selected for use in the field
evaluation test. Supelco's  Carboxen-564,
a hydrophobia carbon molecular  sieve
which is designed  to replace charcoal for
the  collection of  airborne contaminants,
demonstrated  excellent collection
efficiency for all the solvents. After 158
minutes of sampling the solvent mixture
through the Carboxen sorbent tube at a
flow rate  of 0.460 L/min  with  65%
moisture and a stack gas temperature of
173*F, there was  no  evidence  of
breakthrough.  Union Carbide's TRI-X-
100,  which Is an  inorganic  highly
hydrophobia silica oxide  generally used
for waste removal in industrial adsorption,
also showed  acceptable  collection
efficiency.  Under  similar sampling
conditions with 46% moisture, an average
stack gas temperature of  155°F,  and a
sample flow rate of 0.590 L/min, it
showed 20% breakthrough of CCI4, while
the  other three solvents were  retained.
The other five sorbents did not retain  the
solvents under similar conditions.
  The  desorption efficiencies  of the
Carboxen and TRI-X-100 sorbents were
evaluated.  Although  each  of  the
chlorinated solvents could  be analytically
separated  from  each of the desorbing
solutions, methylene chloride proved to
be  the most efficient in  removing  the
solvents from both  sorbents, with the
lowest recovery being 82%. Methylene
chloride  was, therefore,  chosen  for
desorption of the  sorbent tubes for  the
remaining GC analyses.
  GC  analysis   of  the   condensate
collected by the in-line gas conditioning
system for Tedlar bag  sampling
confirmed that the  solvents were  not
being collected; none were detected.
  The  analysis of  samples collected
during the preliminary sampling yielded
an average relative standard deviation of
(RSD) 14.1% and an average accuracy of
 ±9.6%  for  the  solvents collected  on
TRI-X-100. The average RSD for the
solvents collected on the Carboxen-564
was 10.0%, with  an average accuracy of
 ±8.8%.  For the Tedlar bag sample
analysis,  the  precision, in terms  of the
average RSD, was 6.1%, with an average
accuracy of ±9.8%. All adsorption tubes
were desorbed with 4 ml of methylene
chloride.  One problem noted with the
sampling system during the preliminary
test run was that  the sample conditioning
devices  connected to  the  end  of the
manifold farthest  from the stack gas flow
were not collecting  as  much condensate
as the  others. As a result,  the sorbent
tubes  were not equally exposed  to the
high moisture conditions  and system
modification  was necessary to  ensure
that the gas sampled by all the trains was
equivalent. Therefore,  the manifold was
wrapped  with a  heating coil to prevent
loss of heat at the end  of the manifold. It
was also discovered that some  of the
critical orifices were clogging, which was
most likely caused by fine particles of the
sorbent exiting  the tubes. This was
remedied by including filters in-line
behind the tubes.
  The refined sampling system was used
to collect the organic solvents spiked in
the high temperature/high moisture gas-
fired  boiler  emissions.  The overall
average RSD for the small  bag samples
was 13.5%, and the overall average RSD
for the large  bag  samples was 23%, with
an  average  accuracy  of -14.0% and
 + 8.7%, respectively.  For  the sorbent
tube analysis, over  90% of each solvent
was collected on the primary  Carboxen
tube, while as much  as 50%  of each
solvent was found on the backup  TRI-X-
100 tube when the sorbent was exposed
to high  moisture conditions. A pooled
standard  deviation was calculated for
both the TRI-X-100  tubes  and  the
Carboxen  tubes. The  pooled  standard
deviations for TRI-X-100  tubes  were:
4.28%  for CCI4, 12.95%  for CHCI3,
3.13%  for PERC, and  3.06% for TCE.
The pooled  standard  deviations for the
Carboxen  tubes  were: 6.09%  for CCI4,
5.92% for CHCI3, 5.50% for PERC, and
3.18%  for TCE. The  overall  average
standard  deviation for  solvent  collection
on  TRI-X-100 was 5.85%, while the
overall average standard deviation  for the
Carboxen  samples  was 5.17% using  a
combination of the primary and  backup
tube values.  The  average accuracies
calculated for the Carboxen and the TRI-
X-100 tubes were  +8.8% and  +9.8%,
respectively.

Conclusion
  Method 18  sampling  procedures  and
gas  chromatography  using  flame
ionization detection  were found to  be
acceptable for the measurement of PERC
at degreasing facilities.
  Method 18  procedures modified  for
sample  collection   under  high
temperature/high  moisture conditions
were found to be acceptable for sampling
and  analysis  of  emissions  of CCI4,
CHCI3, PERC, and TCE.
  The Supelco sorbent, Carboxen-564, is
effective for sample  collection of CCI4,
CHCI3, PERC, and TCE emissions in high
temperature  and  high  moisture
environments. TRI-X-100 appears to be
effective for  the  collection  of  these
solvents  under   high  temperature
conditions;  however,   its collection
efficiency is decreased in the presence of
high moisture.
  An  in-line gas  conditioning system
employing a condenser  for cooling the
sample gas stream and  a knock-out jar
for eliminating  moisture collection is
appropriate  for Tedlar bag sampling of
the  above  solvents  under high
temperature and-high   moisture  con-
ditions.

References
  1."Method  18:  Measurement  of
    Gaseous  Organic  Compound
    Emissions by Gas Chromatography,"
    40 Code of Federal  Regulations  60,
    July 1, 1987.
  2. Knoll, J. E., M. A. Smith, and M. R.
    Midgett,  "Evaluation of  Emission
    Test Methods  for  Halogenated
    Hydrocarbons"  -   Vol.  1,  CCI4,
    C2H4Cl2 CI2CI4, Publication No. EPA-
    600/4-79-02,  U.  S.  Environmental
    Protection Agency,  Research  Tri-
    angle Park, North Carolina.
  3."Method  18:  Measurement  of
    Gaseous Organic  Compound
    Emissions   by  Gas  Chroma-
    tography," Quality Assurance Hand-
    book for Air  Pollution Measurement
    Systems, Volume III, Publication No.
    EPA-600/4-77-027b, August 1977.

-------

-------
   A.C.  Carver and W.G. DeWees  are with Entropy Environmentalists, Research
        Triangle Park, NC; £ A Coppedge (also the Project Officer) is with the
        Atmospheric Research and Exposure Assessment Laboratory, Research
        Triangle Park, NC. 27711.
   The complete report, entitled "Laboratory and Field Evaluations of Methodology
        for Measuring  Emissions of Chlorinated Solvents  from  Stationary
        Sources," (Order No. PB90-155 565/AS; Cost: $17.00, subject to change)
        wili be available only from:
            National Technical Information Service
            5285 Port Royal Road
            Springfield, VA 22161
            Telephone:  703-487-4650
   The EPA Project Officer can be contacted at:
            Atmospheric Research and Exposure Assessment Laboratory
            U.S. Environmental Protection Agency
            Research Triangle Park, NC 27711
United States
Environmental Protection
Agency
Center for Environmental Research
Information
Cincinnati OH 45268
Official Business
Penalty for Private Use $300
EPA/600/S3-90/009

-------