United States
             Environmental Protection
             Agency
Atmospheric Research and
Exposure Assessment Laboratory
Research Triangle Park, NC 27711
             Research and Development
EPA/600/SR-94/130    May 1995
EPA     Project Summary

             Field  Test of a  Generic
             Method  for  Halogenated
             Hydrocarbons: A VOST Test at  a
             Chemical Manufacturing  Facility
             Using  a Modified VOST
             Sampling Method
             James F. McGaughey, Joan T. Bursey, and Raymond G. Merrill
               Because of the lack of commercial
             availability of SKC Lot 104 petroleum-
             based charcoal, the lot  of charcoal
             specified  for use in the  Volatile Or-
             ganic Sampling Train (VOST, SW-846
             Method 0030), a laboratory study was
             performed to select an appropriate sub-
             stitute. On the basis of performance,
             availability, and cost, Anasorb 747®*
             was selected as the substitute for SKC
             Lot 104, and a draft sampling protocol
             for the modified VOST train was writ-
             ten. In the modified VOST sampling
             train, the  second  tube containing se-
             quential beds of Tenax® and charcoal
             was replaced by a tube containing only
             Anasorb®. This replacement necessi-
             tated the use of two Tenax® front tubes
             in the train, for a total of three sorbent
             tubes. A field evaluation study, using
             dynamic spiking and quadruple sam-
             pling trains, was conducted to provide
             a statistical comparison of the perfor-
             mance of the VOST and the modified
             VOST methodologies for  the volatile
             halogenated organic compounds listed
             in Title III  of the Clean Air Act Amend-
             ments of 1990.
             * Mention of trade names or commercial prod-
              ucts does not constitute endorsement or recom-
              mendation for use.
  This Project Summary was developed
by EPA's Atmospheric Research and
Exposure Assessment Laboratory, Re-
search Triangle Park, NC, to announce
key findings of the research project
that is fully documented in a  separate
report of the same title  (see Project
Report ordering information at back).

Introduction
  Validation of a method for a  particular
analyte or group of analytes means that
the performance of the  methodology for
these analytes has been established and
demonstrated through field tests at the
type of source category of interest: that is,
the precision and bias of the method have
been established experimentally.  Field vali-
dation may be performed by performing
spiking operations in the field so that the
precision and bias  of the method can be
demonstrated from sample collection
through analysis, using either static or dy-
namic spiking procedures as specified in
EPA Method 301, the Validation Protocol.
Another procedure for validation of a
method is a side-by-side comparison of a
candidate method to a validated method
to establish comparable performance for
the same analytes in  the  same matrix
(same source category).

-------
  Under Title  III of the  Clean  Air  Act
Amendments of  1990,  identification  and/
or validation of  sampling and  analytical
methods are required for the volatile halo-
genated organic  compounds  which  are
listed (Table 1). The candidate test method
is  the  Volatile  Organic  Sampling  Train
(VOST), which consists of SW-846 Sam-
pling Method 0030 and  SW-846 Analytical
Method 5040 or  5041.  Method  0030, the
sampling protocol,  requires  the use of
Tenax® and a  specific petroleum-based
charcoal (SKC Lot 104) or its equivalent.
Since SKC Lot 104  charcoal is no longer
commercially available  and Method 0030
contains no performance specifications for
the charcoal, a laboratory study was per-
formed  to identify an acceptable equiva-
lent. Several candidate sorbents were stud-
ied in the laboratory,  and Anasorb 747®
was selected as a candidate sorbent for a
field test because of its performance, avail-
ability, and cost. In the course of the labo-
ratory studies,  the decision was made to
use only Anasorb® in the back tube rather
than the sequential beds of Tenax® and
charcoal which had been used previously.
Use of  a  back tube consisting  of only
Anasorb®  necessitated the  use  of two
Tenax® front tubes in a modified VOST
train. A  new sampling protocol was writ-
ten to incorporate  the use of the modified
sorbent tube structure. The analytical pro-
cedures of Method 5041 were also modi-
fied to include  individual analysis of each
of the  three  sorbent tubes  to  examine
   compound distribution.  In routine applica-
   tion, the analysis would be performed with
   paired Tenax® tubes.
     A presurvey conducted at a chemical
   manufacturing facility established that no
   compounds of interest were present above
   a level of  5 times  the  Method  Detection
   Limit. Evaluation of the  modified  VOST
   method was accomplished by performing
   sampling   and analysis  following  Draft
   Method 0031  and Method  5041,  using
   Method 0030 as the reference method re-
   quired by EPA Method 301.  The bias of
   the method can be evaluated by compar-
   ing the results obtained by one  candidate
   method to the results  obtained  by a vali-
   dated method. To determine  the precision
   of the test method, multiple  or collocated
Table 1.   Volatile Halogenated Organic Compounds,  Title III of the Clean Air Act Amendments, for Which Laboratory Testing has Determined the
          Applicability of the VOST Method
             Compound
                                                       Boiling
                                                        point
                                                                                                 Comments
allyl chloride (3-chloropropene)

bis(chloromethyl) ether

carbon tetrachloride

chlorobenzene

chloroform

chloromethyl methyl ether

chloroprene

1,3-dichloropropylene

epichlorohydrin

ethyl chloride (chloroethane)

ethylene dibromide (1,2-dibromoethane)

ethylene dichloride (1,2-dichloroethane)

ethylidene dichloride (1,1-dichloroethylene)

methyl bromide (bromomethane)

methyl chloride (chloromethane)

methyl chloroform (1,1,1-trichloroethane)

methylene chloride

methyl iodide (iodomethane)

propylene dichloride (1,2-dichloropropane)

tetrachloroethylene

1,1,2-trichloroethane

trichloroethylene

vinyl chloride

vinyl bromide (1,1-dichloroethene)

vinylidene chloride
           44-46

           1061

           77

           1321

           60.5-61.5

           55-57

           59.4

           105-1062

           115-1171

           123

           737-732'

           83

           57
           43

           -24.23

           74-76

           39.8-40

           41-43

           95-96

           727'

           770-775'

           86.9

           -13.43

           1&

           30-32
Acceptable performance in laboratory

Decomposes in water; cannot be analyzed

Recovery too high in laboratory study

Acceptable performance in laboratory

Acceptable performance in laboratory

Decomposes in water; cannot be analyzed

Acceptable performance in laboratory

Acceptable performance in laboratory

Soluble in water; cannot be analyzed

Acceptable performance in laboratory

Acceptable performance in laboratory

Acceptable performance in laboratory

Acceptable performance in laboratory

Recovery unacceptably high in laboratory

Erratic and unacceptably high recovery in laboratory

Recovery too high in laboratory study

Recovery too high in laboratory study

Acceptable performance in laboratory

Acceptable performance in laboratory

Acceptable performance in laboratory

Acceptable performance in laboratory

Acceptable performance in laboratory

Acceptable performance in laboratory

Acceptable performance in laboratory

Acceptable performance in laboratory
'  Above the maximum VOS T boiling point of 100°C; included in the testing because compounds in the range of 100-132°C are frequently tested by the VOS T
 method.
2  Boiling temperature at 730 mm Hg.
3  Below the common lower temperature limit of30°C usually used for VOST.
4  Boiling temperature at 750 mm Hg.

-------
simultaneous samples are taken and ana-
lyzed. In this field test, known concentra-
tions of volatile halogenated analytes were
introduced into the VOST trains and  into
the modified VOST trains, using the same
certified cylinder as a source of the com-
pounds. Using the same dynamic spiking
procedures,  multiple  collocated simulta-
neous  samples  were taken  using  both
methods at the same source.
  Bias is any systematic positive or nega-
tive difference between the measured val-
ues and the true value of a sample. Bias
may result from analytical interferences,
errors  in calibration,  or inefficiencies in
the collection of the  analyte. When the
bias of  the  method  is determined for a
given analyte, a  correction for the bias
may be made. The EPA Method 301 al-
lows for this correction within a range of
90-110% when a candidate method is be-
ing compared to a reference method. Bias
values outside this range may be grounds
for rejecting the candidate method.
  Precision  is the variability in  the data
from the entire measurement system (both
                   sampling and analysis) as determined from
                   multiple or collocated sampling trains. Fol-
                   lowing  the EPA Method 301 procedures,
                   two paired sampling trains are  used  to
                   determine the precision of the entire sys-
                   tem. Use of quadruple (Quad) trains with
                   four collocated  sampling probes allows
                   operation  of two  spiked trains  and two
                   unspiked trains. The EPA Method 301 re-
                   quires that the precision of the candidate
                   method not be significantly  different from
                   the reference method in order for the can-
                   didate  method to be considered accept-
                   able. To determine bias and precision in
                   the field, a total of at  least 16 samples
                   using quadruple collocated sampling trains
                   was collected using both VOST and modi-
                   fied VOST.
                     Statistical  analysis of the results from
                   the field studies with dynamic spiking show
                   that the two  methods are at least equiva-
                   lent, with  the modified VOST methodol-
                   ogy frequently being somewhat  superior
                   in compound recovery and precision.
Procedure
  Sampling was performed by withdraw-
ing  stack  gas from a single  port in  the
stack through a Quad probe, then direct-
ing  the sampled  gas simultaneously to
four similar VOST or modified VOST sam-
pling trains. The Quad probe contains four
similar heated probes that can be inserted
into the stack as one unit. The front end
of the Quad probe was positioned in  the
center  of the  stack and remained in that
location during  each  day of  testing;  no
traverse of the stack was performed with
the  Quad  probe.  The true concentration
of the components of the stack gas was
of no interest to this  program as long as
any quantities of the  compounds of inter-
est  were equal for each train, so travers-
ing  was not required. For modified VOST
sampling,  two of the trains of each Quad
run  were  dynamically spiked  (Trains A
and B) and two were unspiked (Trains C
and D). The configuration of the modified
VOST sampling train is shown in Figure 1.
  Dynamic spiking  of candidate VOST
compounds into the modified VOST train
      Glass
      wool
      filter
                                            Three way
                                           glass/teflon
                                              valve
Spiking gas in
                                       To bubble
                                       flow meter
      Stack
                 Heated glass
                  lined probe
                       (  j    1/4" Teflon
                       v--}      tube
                    Condensate
                       trap
                                                                                        Meter box
                                          Sampling module
Figure 1.  Modified VOST sampling train.

-------
was  performed using a  compressed gas
cylinder containing the volatile halogenated
organic compounds. The  same  cylinder
was  used for spiking both the VOST train
and  the  modified  VOST train,  with  the
composition and concentration of the vola-
tile halogenated organic compounds in the
cylinder established by independent labo-
ratory tests. The  boiling point of 100°C
cited  in Method 0030 was  extended  to
approximately 135°C to allow the inclu-
sion  of chlorobenzene, a compound fre-
quently determined by the VOST method-
ology. The  Modified VOST spiking appa-
ratus shown  in Figure 2, was  used for
dynamic spiking. In the  field, the  spiking
gas was allowed to flow through the spik-
ing apparatus for two hours before direct-
ing the flow to the sampling trains to mini-
mize any adsorptive losses during  actual
Quad sampling runs. A  total of ten (10)
Quad sampling runs for both VOST and
modified VOST was performed in the field
to provide  adequate samples for statisti-
cal comparison of the two methods and to
allow backup samples  in the event that
any samples became invalid due to break-
age or loss of data during analysis.
  Analysis of field samples was performed
according  to  SW-846  Method  5041
(VOST), analyzing each tube individually
to establish distribution of compounds. The
three tubes associated with each modified
VOST  sample set were analyzed in  the
following order:  Anasorb®,  Tenax®  #2,
Tenax®#1. No condensate was observed
in the collection of samples.

Results and Discussion
  Mean recoveries of the volatile haloge-
nated organic compounds using the modi-
fied VOST sampling method are shown in
Table 2. No background corrections were
made because  background levels of the
analytes were less than 5 times the Method
Detection Limit. Methyl chloride exhibited
erratic recoveries, far above 100% in most
analyses. This erratic behavior upon analy-
sis had been observed in  laboratory stud-
ies, in the first  field test of the VOST
method, and  in this field test of both  the
VOST and  modified VOST methods. The
compound is apparently being formed dur-
ing sampling and analysis.
  Surrogate compound  recoveries from
the Tenax® tubes were uniformly high,  so
correction for  surrogate  recoveries was
not made. The same surrogate compounds
were used  to spike the  Anasorb® tubes
prior to analysis, and recovery of the less
volatile  surrogate  compounds from
Anasorb® is poor; correction of Anasorb®
results for surrogate recovery would dis-
tort the analytical results.
  Examination  of the observed distribution
of analytes  among the three sorbent tubes
of the modified VOST sampling train dem-
onstrates the necessity of a second Tenax®
tube  in the sampling train. Methyl  chloride,
vinyl  chloride, and ethyl chloride are found
primarily on the back (Anasorb®) tube. The
following  compounds are found  primarily
(>90%) on the front Tenax® tube: ethylidene
dichloride, chlorobenzene, chloroform, pro-
pylene dichloride,  methyl  chloroform, car-
bon tetrachloride,  ethylene  dichloride,  tri-
chloroethene,   cis-1,3-dichloropropene,
trans-1,3-dichloropropene,  1,1,2-trichloro-
          Regulator
    (temperature controlled)
         Certified
         standard
          (spiking
        compounds)
                             Chassis (temperature controlled)
                                                     To Train A •
                                                                                                        -^- To Train B

                                                                                                        Fine metering
                                                                                                           valves
                                                                                                         (flow control)
                                                                      1/4" teflon line
                                                                     (heat traced and
                                                                        insulated)
                                                                             • ••n Cm a tin a
                                                                                                          Temperature
                                                                                                            display
                                      Variable transformers
                                  (to provide temperature control)
                                       Thermocouple selector
Figure 2.  Modified VOS T spiking apparatus.

-------
Table 2.  Mean Recoveries of Volatile Halogenated Organic Compounds, Using the Modified VOST Method
         Compound
                Mean
               Recovery
               (Percent)1
                                                                                                             Percent
                                                                                                             Relative
                                                                                                             Standard
                                                                                                             Deviation
methyl chloride (chloromethane)
                                                              167.5
                                                                                                              56.4

methyl bromide (bromomethane)
ethyl chloride (chloroethane)
vinyl chloride
vinyl bromide
allyl chloride (3-chloropropene)
45.7
45.3
44.2
38.0
26.0
46.7
30.0
24.2
22.5
21.1
'  The shaded area indicates recoveries from 50 to 150 percent with percent relative standard deviation of 50 or less.
ethane,  tetrachloroethene,  methyl  iodide,
allyl chloride, ethylene dibromide, and chlo-
roprene.  Other  analytes were  distributed
among the three tubes.
  Comparative  results for the VOST and
modified VOST are  shown in Table 3. In
most  instances,  recovery  for modified
VOST is equivalent to  or better than re-
covery for VOST,  and precision for modi-
fied VOST is equivalent to or better than
precision for VOST.
Conclusions and
Recommendations
  Based on  the  field  evaluation  of  the
modified VOST method as compared to
the VOST method,  the following  conclu-
sions may be drawn:

  • On  the basis of a compound-by-com-
    pound  comparison  of recovery and
    precision, the two methods are com-
parable and,  where  they  differ,  the
modified VOST usually shows better
performance with  respect to both re-
covery and precision.
The  only  compound which showed
acceptable performance in the VOST
but not in  the modified  VOST is me-
thyl bromide.  However, the range in
both the  recovery and precision is
sufficiently large that the  differences
between the results for the VOST and
the modified VOST cannot  be consid-
ered statistically significant.
Methyl chloride (chloromethane) has
shown erratic performance, with  ex-
tremely high recoveries and poor pre-
cision in several laboratory studies, in
two field tests  using VOST, and in
the field test using the modified VOST
methodology.  Methyl chloride is  not
an appropriate  analyte for any form
of the VOST method because quanti-
tative calculations  are  unreliable for
this analyte. Even qualitative informa-
tion (i.e., presence or absence) is sus-
pect because methyl chloride may be
formed on the sorbent tubes from de-
composition of other halogenated com-
pounds.
The  following  compounds  were  not
sampled and analyzed successfully in
this field evaluation using either VOST
or modified VOST: methyl chloride
(chloromethane), vinyl chloride, ethyl
chloride (chloroethane), allyl chloride
(3-chloropropene), and vinyl bromide.
Other than methyl chloride, all  of the
analytes which  show  poor  perfor-
mance contain a double bond.  Appli-
cation  of  VOST or modified VOST
methodology for accurate quantitative
determination of analytes containing

-------
Table 3.  Results of VOSTand Modified VOST: Recovery and Precision
                                                    Modified VOST
                                                                                                     VOST
Compound
methyl chloride (chloromethane)
ethylidene dichloride (1,1-dichloroethane)
chlorobenzene
vinyl chloride
vinylidene chloride (1,1-dichloroethene)
chloroform
propylene dichloride (1,2-dichloropropane)
methyl bromide (bromomethane)
ethyl chloride (chloroethane)
methylene chloride
methyl chloroform (1,1,1-trichloroethane)
carbon tetrachloride
ethylene dichloride (1,2-dichloroethane)
trichloroethene
cis- 1 , 3-dichloropropene
trans-1 ,3-dichloropropene
1,1,2-trichloroethane
tetrachloroethene
methyl iodide (iodomethane)
allyl chloride (3-chloropropene)
ethylene dibromide (1,2-dibromoethane)
chloroprene
vinyl bromide
Mean
Percent
Recovery1
167.5
96.2 2
91.62
44.2
96.8 2
98.4 2
149.42
45.7
45.3
120.72
87.1 2
89.3 2
83.2 2
148.72
118.42
75.22
117.32
61. 8 2
89. 02
26.0
108.52
85.8 2
38.0
Percent
Relative
Standard
Deviation
56.4
12.6
13.0
24.2
17.2
20.4
14.0
46.7
30.0
10.9
12.1
12.5
25.1
3.4
21.0
32.6
20.5
8.0
11.9
21.1
23.2
15.3
22.5
Mean
Percent
Recovery1
255.3
86.0 2
84.8 2
37.3
77.8 2
95.3 2
117.72
52. S2
31.4
90.8 2
96.8 2
85. 7 2
78.6 2
124.0 2
83.5 2
47.9
81.42
57.5 2
77.8 2
36.4
81.62
76.4 2
28.4
Percent
Relative
Standard
Deviation
58.1
13.2
27.9
39.5
25.1
14.3
30.0
27.8
37.6
11.7
19.4
13.8
27.7
16.8
16.1
35.0
14.4
12.5
20.4
29.6
31.0
12.3
30.9
'  Mean of 6 runs (12 sets of dynamically spiked tubes).
2  Indicates acceptable performance, using the criteria of recovery from 50 to 150 percent, with percent relative standard deviation of 50 or less.
    a double bond is questionable.  Be-
    cause  recoveries of these analytes
    tend to be  low, there tends  to be a
    low bias in the quantitative determi-
    nation.
  On the basis of the laboratory and field
efforts conducted  for the volatile haloge-
nated organic compounds from Title III of
the Clean Air Act Amendments,  the fol-
lowing recommendations are made:
  • Careful consideration of the chemical
    and physical  properties of candidate
    VOST or modified VOST analytes can
    predict when  successful performance
of the sampling or analytical method-
ology is unlikely. For example, analy-
sis of water-soluble compounds will
either  be completely unsuccessful
or show a   low  bias,  because
water-soluble compounds cannot be
quantitatively  purged from water in
the course of the analysis.
Neither the VOST nor the Modified
VOST  is  recommended  as  a sam-
pling or analytical method for methyl
chloride  (chloromethane). Even  the
accuracy of VOST or modified VOST
as a screening method to  establish
the presence of methyl  chloride in an
emissions matrix is questionable.
Modification or replacement  of the
VOST methodology should be inves-
tigated to provide a sampling and ana-
lytical method feasible for application
to volatile,  polar,  water-soluble ana-
lytes.
If surrogate compound recoveries are
90% or above, no corrections for sur-
rogate recoveries should be made in
either VOST or modified VOST.

-------
   James F. McGaughey, Joan T. Bursey, and Raymond G. Merrill are with Radian
     Corporation, Research Triangle Park, NC 27709.
   Merrill D. Jackson is the EPA Project Officer (see  below).
   The complete report,  entitled "Field Test of a Generic Method for Halogenated
     Hydrocarbons: A VOST Test at a  Chemical Manufacturing Facility Using a
     Modified  VOST Sampling Method," (Order No. PB95-142055;  Cost: $27.00;
     subject to change)  will be available only from
           National Technical Information Service
           5285 Port Royal Road
           Springfield,  VA 22161
           Telephone:  703-487-4650
   The EPA Project Officer can be contacted at
           Atmospheric Research and Exposure Assessment Laboratory
           U.S.  Environmental Protection Agency
           Research Triangle Park, NC 27711
United States
Environmental Protection Agency
Center for Environmental Research Information
Cincinnati, OH 45268

Official Business
Penalty for Private Use $300
      BULK RATE
POSTAGE & FEES PAID
         EPA
   PERMIT No. G-35
EPA/600/SR-94/130

-------