United States
 Environmental Protection
 Agency	
National Exposure
Research Laboratory
Research Triangle Park, NC 27711
 Research and Development
EPA/600/SR-97/035   July 1997
 Project Summary

 Evaluation  of  a Gossen, GmbH
 Model  PAS  1001i  Photoelectric
 Aerosol Sensor for Real-Time
 PAH  Monitoring
Jane C. Chuang and Mukund Ramamurthi
   The objective of this study was to
 evaluate the performance of a Model
 PAS 10001 Photoelectric Aerosol Sen-
 sor (also referred to in this report as
 the "PAH Analyzer" for real-time moni-
 toring of polycyclic aromatic hydrocar-
 bons  (PAH)  in air.  Three tasks were
 conducted in the study: 1) vapor tests,
 2) particle tests, and 3) integrated sam-
 pler comparison tests. In the vapor
 tests, benzene and PAH vapors were
 introduced individually into a 17 m3 en-
 vironmental  chamber as vapors, with
 the chamber air concentrations moni-
 tored by a Trace Atmospheric Gas Ana-
 lyzer (TAG A). The concentrations of the
 spiked chemicals in the air ranged from
 200 ppb for 1-methylnaphthalene to 500
 ppb for benzene. The PAH analyzer did
 not yield any response when sampling
 these chemical vapors.
   In the particle tests, sodium chloride
 aerosols of  various  sub-micron sizes
 (0.05-0.5 um in  diameter) were gener-
 ated and then sampled by the PAH ana-
 lyzer to determine whether the instru-
 ment responded only to aerosols con-
 taining PAH. The results showed that
 the PAS did respond to NaCl aerosol
 that was anticipated to contain no PAH.
 This response depended on  the size
 and number concentration, and was a
 small fraction of the typical response
 obtained at particle sizes and concen-
 tration levels common in indoor air (for
 example, 0.1-0.3 um diameter, 10,000-
 50,000 particles/cm3). Further step-by-
 step tests confirmed that the analyzer's
 response did result from the passage
 of the aerosol through the ultraviolet
 radiation cell. The origin of this re-
 sponse could not be resolved in this
 study, but possible causes of the re-
 sponse to non-PAH  aerosol were ex-
 plored.
   In the integrated sampler compari-
 son study, two Battelle-developed in-
 door PAH samplers were  collocated
 with the PAH analyzer in indoor envi-
 ronments with and without the  pres-
 ence of environmental tobacco smoke
 (ETS). The fine particle-bound (<2.5 UJTI)
 PAH concentration in the air sampled
 was then estimated by summing the
 fine particle concentrations of individual
 PAH and alkylated PAH species deter-
 mined by gas  chromatography-mass
 spectrometric (GC/MS) analysis of the
 filters and XAD-2 sorbent traps from
 the integrated samplers. The response
 of the PAH analyzer (in pAmp) over the
 corresponding sampling  periods was
 also averaged and then converted to a
 PAH  concentration   using   the
 manufacturer's suggested conversion
 factor of 3000 ng/m3/pAmp.
   In the presence of ETS, the PAH ana-
 lyzer concentrations  were approxi-
 mately 4 times higher than the fine par-
 ticle PAH concentrations derived from
 the integrated samplers. In the absence
 of ETS, the  PAH analyzer  concentra-
 tion was similar to the fine particle-
 bound PAH concentration in one test,
 but was only one-third of the fine par-
 ticle PAH in the other test.  It was also
 observed from the tests conducted that
 the PAH analyzer yielded a larger unit
 response to PAH in the ETS aerosol
 than to PAH in the non-ETS indoor aero-
 sol.
    This Project Summary was developed
 by EPA's National Exposure Research
 Laboratory, Research Triangle Park, NC,
                                                 Printed on Recycled Paper

-------
 to announce key findings of the re-
 search project that is fully documented
 In a separate report of the same title
 (see Project Report ordering informa-
 tion at back).

 Introduction
   The  PAH analyzer evaluated  in this
 study was a Model PAS 1000i Photoelec-
 tric Aerosol  Sensor  manufactured by
 Gossen, GmbH (Erlangen, Germany) and
 obtained  in the U.S. with  ancillary data
 acquisition software from EcoChem Tech-
 nologies,  Inc.  (West Hills, CA).
   The PAH analyzer is based on the prin-
 ciple of photoelectric aerosol charging re-
 ported in  the literature within the last de-
 cade. The presence  of  a mono- or
 submono-layer  of  PAH compounds
 adsorbed on the surface of aerosol par-
 ticles allows UV-light induced photoemis-
 sion of electrons from electron-rich PAH
 molecules that are adsorbed onto a mono-
 layer on  the  surface of a particle. This
 phenomenon,  also referred to as aerosol
 photoemission (APE), is the measurement
 principle used in the PAH analyzer evalu-
 ated in this study.
   The aerosol sampled enters  the ana-
 lyzer via  an electrofilter. The electrofilter
 in the PAS 1000! has a nominal operating
 voltage potential of =465 V, and removes
 all the ions and a fraction of the charged
 particles that  may be present in the air
 being sampled. The aerosol  exiting the
 electrofilter enters  the UV irradiation cell
 where it is irradiated and APE occurs. In
 principle, APE results from the  PAH-
 adsorbed  aerosols  at particle sizes <1 urn.
 The net APE signal is minimal at particle
 sizes >1  urn in diameter; at these sizes
 recapture of the photo-emitted electrons
 by the particles is believed to occur too
 rapidly to  permit the charged particles to
 be detected. The photoemission of elec-
 trons by PAH adsorbed on the surfaces of
 aerosol particles causes the particles them-
 selves to become positively charged. The
 charged aerosol is then brought through
 an electrically isolated conduit to an elec-
 trometer.  The current measured by the
 electrometer is then a sensitive  indicator
 of the degree of APE occurring in the UV
 irradiation cell.
   Previous evaluation studies reported in
 the literature have reported that APE arises
 only from  PAH adsorbed on aerosols. At
the UV irradiation energy used of ~4.2 eV
 no other adsorbate was found  in these
studies to  result in an APE signal in these
 laboratory APE systems. In addition, the
APE signal was found to be proportional
to the amount of adsorbed PAH, up to the
limit of a monolayer on the surface of a
 particle. Researchers have suggested,
 however, that in typical  residential com-
 bustion exhausts and except for residen-
 tial wood combustion, the particle number
 concentrations emitted are sufficiently high
 in relation  to the PAH mass emitted that
 only low PAH surface coverages are likely
 to result. Under these circumstances, the
 net APE response could continue to be
 proportional to the total adsorbed PAH.
   In addition to  these characteristics, the
 APE signal from the analyzer varies with
 the PAH compound adsorbed on the par-
 ticle surface, with the highest  response
 associated with large n- electron systems.
 Of the PAH compounds tested, coronene
 was the most photoelectrically active. The
 nature of the surface on which the PAH
 compounds are  adsorbed also affects the
 APE response.  For each type of particle
 surface, however, the response generally
 increases linearly with increasing surface-
 adsorbed PAH mass.
   Various studies with laboratory APE sys-
 tems have shown that for specific sources
 of PAH aerosols, such as cigarette smoke,
 oil stoves, spark-ignition engines, etc., the
 APE signal is reasonably correlated with
 the concentration of particle-phase PAH
 compounds measured by  wet-chemical
 techniques. This result is consistent with
 the two  characteristic response features
 described above, since the mix of PAH
 compounds and  the aerosol surface char-
 acteristics would tend to remain generally
 constant for a specific source of particu-
 late-phase PAH.
  The APE signal can be expected to be
 less correlated with the particle-phase PAH
 concentration when the mix of PAH com-
 pounds and the aerosol  surface charac-
 teristics in the air sampled are more vari-
 able. This situation  is of significant inter-
 est in evaluating the PAH analyzer as a
 reliable indoor screening or monitoring tool
 since indoor air  can conceivably contain
 particulate-phase PAH originating in vary-
 ing proportions from many different aero-
 sol sources such as ETS, woodstoves,
 burners, fireplaces,  combustion-derived
 aerosols in outdoor air, etc. An evaluation
 of the  relationship  between  the PAH
 analyzer's response and  measured PAH
 in different indoor environments has, how-
 ever, not been previously  conducted.
  The objective of this study was to evalu-
ate the overall performance of the PAH
analyzer in monitoring PAH indoors, with
the following three tasks:

  (1)  Vapor tests to verify that the PAH
      analyzer does not respond to PAH
      vapors,  as reported  previously in
      the literature.
  (2)  Particle tests to determine the re-
      sponse of the PAH analyzer to so-
      dium chloride (NaCI) test aerosols
      and to determine whether the PAH
      analyzer has the desired selective
      response for only PAH containing
      aerosols.
  (3)  Integrated sampler comparison
      tests to compare  the indoor PAH
      concentrations determined  from
      conventional  integrated sampling
      and subsequent  GC/MS  analyses
      with the  average responses over
      corresponding sampling periods
      from the collocated PAH analyzer.

Procedure
  A brief description of the procedures
used to conduct the three tasks is  pro-
vided here:
  (1)  Benzene and PAH vapor tests: Tests
      of the  PAH analyzer's response to
      benzene, p-dichlorobenzene, naph-
      thalene, 1-methylnaphthalene, quino-
      line,  1-chloronaphthalene, and in-
      dene were conducted by introducing
      the vapors individually into a 17 m3
      environmental chamber  and moni-
      toring  the chamber  air  concentra-
      tions by a Trace Atmospheric Gas
      Analyzer (TAGA). The  concentra-
      tions of the spiked chemicals in the
      chamber air ranged from 200  ppb
      for 1-methylnaphthalene to 500 ppb
      for benzene. An aliquot of each of
      the chemicals was injected through
      a heated injection port where it was
      vaporized, and subsequently  dis-
      persed into the chamber. Prior to
      injection, the chamber was flushed
      with AADCO zero air. During the
      experiment, the  chamber  was
      sampled  by  the TAGA, the PAH
      analyzer, and by a TSI Model 3020
      condensation  nucleus  counter
      (CNC)  for monitoring the particle
      concentration in the  chamber. The
      measured concentration  of  the
      spiked vapor was recorded for each
      test, together with the humidity, tem-
      perature, and particle concentration
      in the  chamber atmosphere,  and
      the response of the  PAH analyzer
      when sampling from  the chamber.

  (2)  Particle tests: NaCI test aerosols in
      the 0.05-0.5 nm  size range were
      generated using  a  nebulizer that
      generates a droplet spray by aspi-
      rating  a solution  of NaCI from  a
      reservoir  (analytical  grade NaCI,
      muffled at 400°C for 4 hours was
      used to prepare the salt solution).

-------
   The nebulized droplets were dried
   either by heating in a coiled copper
   tube maintained at ~150°C or by
   dilution with dry air.  The solid salt
   particles were  then  brought to a
   state of charge equilibrium by pas-
   sage  through a radioactive  Kr-85
   neutralizer (2 mCi, 5/89). The neu-
   tralized aerosol was mixed with dry,
   filtered air and introduced  into a
   Plexiglas aerosol  mixing  chamber
   from which it was sampled  by  the
   PAH analyzer and two aerosol mea-
   surement instruments, a  TSI, Inc.
   Model 3020 CNC  and a PMS, Inc.
   Model LAS-X Laser  Aerosol Spec-
   trometer. Size distributions of  the
   test  aerosols were  measured by
   coupling a TSI Model 3040/3042
   Diffusion Battery with the CNC.

   Apart from the NaCI aerosol tests,
   deionized water without dissolved
   salt was  nebulized, dried, and
   sampled from the  aerosol chamber
   to establish background conditions.
   Carrier air alone was also sampled
   from the aerosol chamber in other
   background tests. A second series
   of detailed PAH analyzer tests were
   also conducted in which  two spe-
   cific components  of the  analyzer,
   namely the electrofilter and the UV
   lamp were turned  off selectively
   without disabling  the electrometer.
   These detailed tests of the response
   of the PAH  analyzer to NaCI  test
   aerosols were  conducted using  a
   660-1 stainless steel chamber.  The
   large chamber allowed the  test
   aerosols to be aged for  several
   hours to simulate a typical  indoor
   aerosol before being sampled by
   the PAH analyzer.
(3) Integrated sampler  comparison
   tests: The integrated  sampler  sys-
   tem  consisted of  two  sampling
   trains. The first used a Battelle-de-
   veloped indoor sampler  equipped
   with  an open-face  47-mm  quartz
   fiber filter upstream of an XAD-2
   resin trap and collected the com-
   bined vapor  and  particle-phase
    PAH. The second  train  consisted
    of the following components (listed
    in the order of air flow), impactor-
    denuder-quartz fiber filter-XAD trap,
    and  was designed to collect fine
    particle-bound  (<2.5  ^m)  PAH  in
    the sampled air stream.  The sam-
    pling flow rate through the two sam-
    pling trains was controlled at  20.0
    ± 0.2  Ipm  using a  pump  and a
    metering valve and was  monitored
     continuously by an in-line mass flow
     meter.
     The filter and XAD-2 trap from each
     integrated sampler were combined
     and extracted with dichloromethane
     (DCM) by the Soxhlet technique.
     The DCM extract was concentrated
     by Kuderna-Danish (K-D) evapora-
     tion. The extract was then analyzed
     by electron impact (El) GC/MS for
     target PAH and alkylated PAH spe-
     cies.
     Four  sets of indoor integrated
     samples were  collected during the
     study: two tests were  performed in
     the presence of ETS and two tests
     were  conducted  without ETS. The
     sampling periods employed in the
     tests varied between 8 and 24 hours
     depending on the presence or ab-
     sence of  ETS.  During  each test,
     the PAH  analyzer was collocated
     with the two sampling  trains and
     the response over the sampling pe-
     riod was continuously recorded and
     stored in  a data acquisition com-
     puter. During all  tests, a TSI Model
     3020 CNC was used to monitor the
     particle concentrations in the indoor
     environment being sampled. The
     particle number concentration mea-
     sured by the CNC was recorded by
     a strip-chart recorder that also  re-
     corded the PAH  analyzer response
     during the sampling intervals.

Results
  Benzene and PAH  Vapor Tests. The
PAH analyzer did not yield a response
when sampling chamber air containing
each of the various chemical vapors. This
finding  agrees with the  published litera-
ture  reports that the PAH analyzer does
not respond to vapor-phase  PAH  com-
pounds. The series of seven chamber tests
were conducted at  low humidity  condi-
tions (9% relative humidity at 23-25°C),
and  particle concentrations  in the  cham-
ber remained below 1-2  particles/cm3 in
all tests, i.e., there was minimal possibility
for water vapor-induced hydrolysis and/or
condensation processes that might result
in particles with surface-adsorbed  PAH.
Additional vapor tests were conducted sub-
sequently at chamber humidity conditions
of ~80%, using similar vapor concentra-
tions and temperature conditions.  These
tests also showed no PAH analyzer re-
sponse to vapor-phase PAH in all  except
one  case.  In the one case where a re-
sponse was observed, for indene, a sub-
stantial particle concentration of ~50,000
particles/cm3 was measured in the cham-
ber suggesting that the  observed PAH
analyzer response was due to the adsorp-
tion of vapor-phase PAH onto particle sur-
faces. The particle formation observed was
traced to the polymerization of the aged
indene liquid used in the test (Merck In-
dex). A  more detailed discussion of these
tests will be reported separately.
  Particle Tests. The  PAH analyzer did
not show any response during the back-
ground  tests, when sampling either dry,
filtered  air or filtered  air carrying dried,
deionized  water droplets  from the nebu-
lizer.  The  PAH analyzer did, however,
show a  response when sampling the three
different NaCI test aerosols generated at
various  number concentrations in the aero-
sol chamber (the test aerosols had log-
normal  size  distributions and geometric
mean diameters of ~0.04, -0.08, and -0.15
|im). The non-zero PAH analyzer response
increased approximately linearly with par-
ticle concentration and with increasing test
particle  size. More detailed tests conducted
subsequently also established that the re-
sponse observed was definitely a result of
the passage of the test  aerosol through
the UV irradiation cell. The magnitude of
the response observed for NaCI particles
is generally smaller than that  observed in
either  typical  indoor air  containing low,
background amounts of PAH-coated par-
ticles or high concentrations  of cigarette
smoke  particles.
   An analysis of the PAH content of the
muffled bulk NaCI used  in preparing the
salt solutions was conducted using DCM
extraction and GC/MS analysis. PAH was
not detected in the bulk  salt extract, but
trace amounts of  aliphatic hydrocarbons
and phthalates were detected. It  is pos-
sible that the observed signal is related to
other unknown organic species, and this
possibility could be examined further in a
future study.
   While the results obtained  in this study
 appear to contradict literature reports by
 other researchers, recent tests in Novem-
 ber 1992 by Burtscher and co-workers in
 ETH, Zurich (a research  group closely in-
 volved  with the development of the Gossen
 PAS) using a laboratory APE system also
 found  APE response to nebulized  NaCI
 tests aerosols. These responses  were
 found  to be  of a similar magnitude as .
 those  found  in  this  study. Further,
 Burtscher et al. found that APE is  elimi-
 nated by using an evaporation-condensa-
 tion NaCI aerosol generator that involves
 heating the bulk NaCI to  a temperature of
 ~700°C, suggesting that trace organic con-
 tamination may indeed be responsible for
 the observed APE response from nebu-
 lized NaCI aerosols.

-------
  Integrated Sampler Comparison Tests.
The integrated sampler comparison tests
were used to conduct comparisons  be-
tween the vapor+particle-phase PAH and
fine particle-bound PAH concentrations
from the integrated samplers and the PAH
analyzer's average response (in pAmp)
over the corresponding sampling interval.
The fine particle-bound  (< 2.5 \im) PAH
concentration was estimated by summing
the fine particle concentrations of individual
PAH  and alkylated PAH species deter-
mined from  GC/MS  analysis. PAH con-
centrations were calculated from the aver-
age PAH analyzer response for each cor-
responding test using the conversion fac-
tor  of 3,000 ng/m3/pAmp  provided by
Gossen, GmbH.
  The average  PAH concentrations from
the PAH  analyzer were about 4 times
higher than the measured fine particle PAH
concentrations for the two tests conducted
in the presence of ETS. In the absence of
ETS, the average PAH concentration from
the PAH analyzer was similar to the mea-
sured fine particle PAH concentration from
the integrated sampler in one test but was
one-third of the  fine particle  PAH concen-
tration in the other test.
  Conversion factors were also  derived
on the basis of the fine particle PAH con-
centrations and the PAH analyzer's aver-
age response in each of the four tests.
The calculated  conversion  factors  were
similar for the two tests  within the ETS
category, when  fine particle  PAH was de-
fined  either as the sum of all species, or
the sum of ^ 3-ring species. Within  the
non-ETS category, the calculated conver-
sion factors from one test were signifi-
cantly higher than the conversion factors
calculated from  the data  collected in  the
other test. The conversion factors derived
from the tests were generally within (with
the exception of the one non-ETS test)
the range of the various  instrument cali-
bration  functions  provided   by Gossen,
GmbH.
  Another observation that was also evi-
dent from the non-ETS and ETS tests is that
the  instrument response (in pAmp) was con-
siderably higher when sampling  cigarette
smoke aerosols than it was to the general
indoor aerosol. This result is consistent with
reports in the literature that APE is the highest
(per unit mass) from PAH such as coronene,
benzo(g,h,i)perylene, benzo(e)pyrene,  and
benzo(a)pyrene,  all of which are species
present in ETS aerosol. Another possible rea-
son for this higher  response rate for ETS
aerosol is that a greater fraction of the PAH in
these freshly generated aerosols may be
present in a  surface-adsorbed state  com-
pared to the typical, aged non-ETS indoor
aerosol. There may also be differences in
the surface  characteristics of ETS aero-
sols that result in a higher response rate
compared with non-ETS indoor aerosol.

Conclusions  and
Recommendations
  The following conclusions can be drawn
from this study:
  (1)  The PAH analyzer does not yield a
      response when sampling air from a
      test chamber  atmosphere spiked
      with benzene and PAH vapors. In
      these tests, the aerosol concentra-
      tion in the chamber remained  be-
      low 10 particles/cm3, thus minimiz-
      ing the opportunity  for APE from
      PAH vapors adsorbed on particle
      surfaces.

  (2)  The PAH  analyzer did show a re-
      sponse to pure NaCI aerosol. The
      exact mechanism for this response
      is presently uncertain; one possible
      origin of the observed response is
      APE from trace organics (other than
      PAH)  in the bulk NaCI used to gen-
      erate  the aerosol. The high-purity,
      reagent-grade NaCI used was, how-
      ever,  muffled for 4 hours in a fur-
      nace at 400°C prior to use in  the
      experiments.

  (3)  In  the  presence  of ETS, the PAH
      concentrations calculated from  the
      PAH analyzer's average responses
      were approximately four times (4X)
      higher than the fine particle (<  2.5
      (im) PAH  concentrations derived
      from  integrated  sampler-GC/MS
      measurements. In the absence of
      ETS, the calculated concentration
      from the PAH analyzer's average
      response was similar to the fine
      particle PAH concentration in one
      test but was only one-third  of the
      fine particle  PAH concentration in
      the other  test. The average  re-
      sponses of the analyzer were con-
      verted  to  PAH  concentrations  for
      these  comparisons  using  the
      manufacturer's suggested calibra-
      tion constant of  3,000  ng/m3 per
      pAmp of analyzer signal. It was also
      observed from the tests  that the
      analyzer yielded  a larger unit  re-
      sponse to PAH in ETS than to PAH
      in non-ETS indoor aerosol.
  (4)  The PAH analyzer provided a real-
      time (< 5 sec) response that was
      proportional  to indoor particulate-
      phase  PAH.  The response factors
      relating instrument signal to actual
      fine particle PAH  concentration ap-
      pear to vary with the nature of the
      indoor aerosol being sampled.
  Based on  these  results, the following
major recommendations were developed:
  (1)  The origin of the observed response
      of the PAH analyzer  to  non-PAH
      aerosols is of considerable impor-
      tance  and  must hence be under-
      stood. A plausible explanation for
      this phenomenon is that of APE
      from trace species, such as organ-
      ics, other than PAH. However,  the
      identity of these species is not clear
      at the  present time and hence it is
      difficult to speculate on the possi-
      bility of APE  from these or similar
      organic species adsorbed  on indoor
      aerosols. Several tests to determine
      if the  response  is indeed due to
      trace organic species could be con-
      ducted, and the identity  of these
      species could be isolated.

      If the  response  of the PAH ana-
      lyzer  to  non-PAH  aerosols is  in-
      deed due to the trace organic spe-
      cies, the extent to which these spe-
      cies  would be present in  indoor
      aerosol must be determined. If pos-
      sible,  the analyzer's  background
      response when sampling different
      types  of indoor environments must
      be estimated so that  background
      corrections  can be made depend-
      ing on the particle number concen-
      tration  in the air sampled.
  (2)  Another  issue that must be  ad-
      dressed is the need and function of
      the low-voltage electrofilter used in
      the PAH analyzer.  In  the current
      operating mode,  the electrofilter is
      removing gaseous ions as well as
      a small fraction of the charged aero-
      sol. However, removing a portion
      of the  charged fraction of  the
      sampled aerosol may result in the
      removal of a  possibly significant
      portion of the particulate PAH in
      the air. This effect may be signifi-
      cant for combustion aerosols that
      may be highly charged  immediately
      after production. On the other hand,
      sampling an  initially positively  or
      negatively charged  aerosol could
      interfere with  the measurement of
      APE from adsorbed PAH. The need
      for the  electrofilter must  thus  be
      studied and the impact  of removing
      the electrofilter on instrument per-
      formance investigated.

  (3)  The application of the PAH ana--
      lyzer to indoor air can be addressed
      once the above issues  have been

-------
investigated and resolved. A series          determined from such tests would          combustion sources and in differ-
of comparison tests between  inte-          be beneficial to the potential appli-          ent  indoor environments. We also
grated samplers and the PAH  ana-          cation of the  analyzer as an indoor          recommend investigating the appli-
lyzer can be conducted to correlate          air screening  or monitoring tool. The          cation of the PAH analyzer for moni-
the analyzer response to PAH  con-          PAH  analyzer could also be used          taring indoor particulate-phase  ni-
centration under various indoor air          to measure PAH aerosol size distri-          tro-PAH.
conditions. The response factors          butions, both for the most common

-------

-------

-------
   Jane C. Chuang and Mukund Ramamurthi are with Battelle, Columbus, OH 43201-
     2693.
   Nancy K. Wilson is the EPA Project Officer (see below).
   The complete report, entitled "Evaluation of a Gossen, GmbH Model PAS 100H
     Photoelectric Aerosol Sensor for Real-Time PAH Monitoring," (Order No. PB97-
     147938; Cost: $21.50, subject to change) will be available only from:
           National Technical Information Service  •
           5285 Port Royal Road
           Springfield, VA 22161
           Telephone: 703-487-4650
   The EPA Project Officer can be contacted at:
           National Exposure Research Laboratory
           U.S. Environmental Protection Agency
           Research Triangle Park, NC 27711
United States
Environmental Protection Agency
Center for Environmental Research Information
Cincinnati, OH 45268
Official Business
Penalty for Private Use $300
     BULK RATE
POSTAGE & FEES PAID
         EPA
   PERMIT NO. G-35
EPA/600/SR-97/035

-------