United States        Office of Research and   EPA/600/R-94/038b
         Environmental Protection    Development      April 1994
         Agency          Washington, DC 20460
&EPA   Quality Assurance
         Handbook for
         Air Pollution
         Measurement
         Systems

         Volume II: Ambient Air
         Specific Methods
         (Interim Edition)

-------

-------
                                            EPA-600/R-94/0385
          QUALITY ASSURANCE HANDBOOK

                          FOR

      AIR POLLUTION MEASUREMENT SYSTEMS
Volume II ~  Ambient Air Specific Methods
                 (Interim Edition)
          U.S. ENVIRONMENTAL PROTECTION AGENCY
             OFFICE OF RESEARCH and DEVELOPMENT
 ATMOSPHERIC RESEARCH and ENVIRONMENTAL ASSESSMENT LABORATORY
         RESEARCH TRIANGLE PARK, NORTH CAROLINA 27711
                                        •;X A'.' Printed on Recycled Paper

-------

-------
           OVERVIEW OF THE  INTERIM EDITION OF VOLUME II

      The  Quality Assurance (QA)  Handbook is  comprised of five
 volumes:   Volume I (Principles),  Volume  II (Ambient Air Methods),
 Volume  III (Stationary Source  Methods),  Volume  IV (Meteorological
 Measurement's), and Volume V (Precipitation Measurement Systems) .
 Much  of the material  in Volumes  II,  III  and  V are out-of-date and
 some  portions  of these volumes have  long been out-of-print.

      EPA  is now  preparing  an updated version of the QA Handbook
 series  which will be  available in September  1995.   To meet the
 needs of  the user community until the  updated version is
 available, EPA has published Interim Editions of Volumes I',  II,
 III,  IV and V.   Each  volume of the Interim Editions,  is being
 issued  as a complete  unit  with out-of-date sections either
 deleted or modified using  addendum sheets  and handwritten
 notations in the text.

      This volume and  the-other four  volumes  of  the  Interim
 Edition of the QA Handbook are available  at  no  charge from-
               USEPA/ORD
               Center for  Environmental Research Information
               26  West  Martin  Luther King  Drive
               Cincinnati, Ohio  .45268

      For  the reasons  given below  the following  six  sections
 published in the original  edition  of Volume  II were  excluded from
 this'edition of  the handbook.      .  •      "    .

      Section 2.0.7  (Protocol 2 Gases) was  combined'with Section
 3.0.4 of  Volume  III (Protocol  1 Gases)  and published  as  a
 separate  document  entitled "EPA Traceability  Protocol  for Assay
 and Certification  of  Gaseous Calibration Standards  (Revised
 September  1993),"  EPA 600/R93/224).  This document provides
 guidance  to those  who prepare  and  sell gaseous pollutant
 standards  traceable to  the National  Institute of Standards' and
 Technology (NIST).

      Section 2.0.8  (Calculating Precision and Accuracy  for SLAM
 and PSD Analyzers) was deleted because it has been supplanted by
 guidance available in Part 58 of Title 40 of the Code of Federal
 Regulations (40  CFR 58, Appendix A).  Section 2,1.(SO2 by   '  -
 Pararosaniline)  and Section 2.4  (NO2  by Sodium Arsenite) were
 deleted because  these methods are no longer used in the United
 States  (U.S.).    Section 2.5 (SO2  by Flame Photometry)  and
 Section 2.7 (O3 by chemiluminescence) were deleted because they
 are used by less than 1 percent of the U.S. organizations
measuring air pollution.

     These six deleted sections can be obtained by writing to-
          QA Handbook Coordinator
          US EPA/ORD/AREAL/MD 77B
          Research Triangle Park, NC  27711

-------
      Many of  the EPA contacts  and organizational units identified
 in Volume II  are no longer correct and some  of the reference
 materials and procedures  cited have been discontinued or
 replaced.   This  type of put-of-date information is widely
 dispersed throughout Volume II.   Rather than change every
 affected section,  for clarity  and neatness sake,  we have provided
 below a  listing  of the original  information  and the corresponding
 updated  information.
      1)   NBS  is  now the National Institute of Standards and
 Technology (NIST)1
      2)   EMSL is ,now the  Atmospheric Research and Exposure
 Assessment Laboratory (AREAL).  •
      3)   QAD  is  now the Quality  Assurance and Technical Support
 Division (QATSD/AREAL).
      4)   CRM  is  now NIST  Traceable Reference Material (NTRM).
      5)   SAROAD  is now the Aerometrie Information Retrieval
 System (AIRS).
      6).   The  National Aerometrie Data Bank  (NADP)  and the SLAMS
 Precision and Accuracy Reporting System (PARS)  are now included
 in AIRS.
      7)   The  TSP (total suspended particulate matter)  standard
 has now  been  replaced with a standard based  on particle size  (PM-
 10) .
      8)   Correspondingly  the TSP sample (hi  vol)  has  been
 replaced with size  selective  (PM-10)  particle samplers.
      9)   The  address  to obtain the  most .recent  "List  of EPA
 Designated Reference  and  Equivalent  Methods"  is now:
           US  EPA/ORD/AREAL/MRDD/MD-77
           Research  Triangle Park, NC  27711
      10)   Due to the  current widespread use  of  data loggers
 references  made  to  chart  recorders  should also  be  considered to
 include  data  loggers.

      In  the updated version of Volume  II, which wil-1  be  available
 in  September  1995,  these  changes  will  be  included  in  the  text.
 Sections on QA for  organic compound  measurement systems,  for O3
 by UV and  for automated PM-10 samplers  will  be  added  and  a  new
page numbering system will be used.

    • Finally,  the user of  the' QA  Handbook is  cautioned  to bear in
mind that  the information provided in  the handbook is"for
guidance purposes only.  EPA regulations  are published  in the
Code of  Federal  Regulations (CFR).  When  information  in  the CFR
conflicts with information in the QA Handbook, the CFR  shall be
considered the authoritative and  legally bonding document.

                                   William J. Mitchell
                                         Chief
                              Quality Assurance Support Branch

-------
                       Volume II
                  (Interim Edition)

                 Table of  Contents
       Overview of the Interim Edition of Volume II
2.0    General Aspects of Quality Assurance for
       Ambient Air Monitoring Systems

2.0.1   Sampling Network Design and Site Selection
2.0.2  Sampling Considerations
2.0.3  Data Handling and Reporting
2.0.4  Reference and Equivalent Methods
2.0.5  Recommended Quality Assurance Program
       for Ambient Air Measurements
2.0.6  Chain-of-Custody Procedures for Ambient
       Air Samples
2.0.7—Traccability Protocol for Establishing True
     .  Concentrations of Gases
       Uocd for Calibration and Auditn of Air
       Pollution Analyzers (Protocol No. 2)
2.0.8—Calculationo to Aosooo Monitoring Data for
       Prccioion  ond Accuracy for SLAMS ond PSD
       Automated Analyzers and Monuol Mcthoda
2.0.9  Specific Guidance for a Quality Control
       Program for SLAMS and  PSD Automated
       Analyzers and Manual Methods
2.0.10 USEPA National Performance Audit Program
2.0.11  System Audit Criteria and Procedures for
       Ambient Air Monitoring Programs
2.0.12 Audit Procedures for Use by State and Local
       Air Monitoring Agencies
       Reference! Method for the Determination of
       Sulfur Dioxide in the Atmoophcrc
       (Pororosonilinc Method)
2.2    Reference Method for the Determination of
       Suspended Particulates in the Atmosphere
       (High-Volume Method)

2.2.1   Procurement of Equipment and Supplies
2.2.2   Calibration of Equipment

-------
  2.2.3   Filter Selection and Preparation
  2.2.4   Sampling Procedure
  2.2.5   Analysis of Samples
  2.2.6   Calculations and Data Reporting
  2.2.7   Maintenance
  2.2.8   Auditing Procedure
  2.2.9   Assessment of Monitoring Data for Precision
         and Accuracy
  2.2.10  Recommended Standards for Establishing
         Traceability
  2.2.11  Reference Method
•  2.2.12  Rcfcrcncca
  2.3.13  Data forma

  2.3     Reference Method for the Determination of
         Nitrogen Dioxide in the Atmosphere
         (Chemiluminescence)

  2.3.1    Procurement of Apparatus and Supplies
  2.3.2   Calibration of Equipment
 2.3.3   Operation and Procedure
 2.3.4  Data Reduction, Validation and Reporting
 2.3.5  Maintenance
 2.3.6  Auditing Procedure
 2.3.7  Assessment of Monitoring Data for Precision
        and Accuracy
 2.3.8  Recommended Standards for Establishing
        Traceability
 2.3.0—Reference Method
 2.3.10 References
 2.3.11  Data forma

 3r4	Equivalent Method for the Determination of
        Nitrogen Dioxide in the Atmosphere
        (Sodium Arocnitc)
        Equivalent Method for the Determination of
        Sulfur Dioxide in the Atmosphere (Flame
        Photometric Detector)
 2.6     Reference Method for the Determination of
        Carbon Monoxide in the Atmosphere
        (Non-Dispersive Infrared Spectrometry)

 2.6.1   Procurement of Equipment and Supplies
 2.6.2   Calibration of Equipment
 2.6.3   Operation and Procedure
 2.6.4   Data Reduction, Validation and Reporting
 2.6.5   Maintenance
 2.6.6   Auditing Procedure
 2.6.7   Assessment of Monitoring Data for Precision
        and Accuracy

-------
 2.6.8  Recommended Standards for Establishing
        Traceability

 2r7	Reference Method for the Determination  of
        Ozone in the Atmosphere (Chcmilumincoccncc)
 2.8    Reference Method for the Determination of
        Lead in Suspended Particulate Matter
        Collected from Ambient Air (Atomic
        Absorption Spectrometry)

 2.8.1   Procurement of Equipment and Supplies
 2.8.2   Calibration of Equipment
 2.8.3   Filter Selection'and Procedure
 2.8.4   Sampling Procedure
 2.8.5   Analysis of Samples
 2.8.6   Calculations and Data Reporting
 2.8.7   Maintenance
 2.8.8   Auditing Procedure
 2.8.9   Assessment of Monitoring Data for Precision
        and Accuracy
 2.8.10  Recommended Standards for Establishing
        Traceability

 2.9     Reference Method for the Determination of
        Sulfur Dioxide in the Atmosphere'
        (Fluorescence)

 2.9.1   Procurement of Apparatus and Supplies
 2.9.2   Calibration of Equipment
 2.9.3   Operation and Procedure
 2.9.4   Data Reduction. Validation and Reporting
 2.9.5   Maintenance
 2.9.6   Auditing Procedure
 2.9.7   Assessment of Monitoring Data for Precision
        and Accuracy
 2.9.8   Recommended Standards for  Establishing
        Traceability

 2.10    Reference Method for the Determination of Particulate
        Matter as PM10 in the Atmosphere (Dichotomous
        Sampler Method)

 2.10.1  Procurement of Equipment and Supplies
 2.10.2  Calibration Procedures
 2.10.3  Field Operations
 2.10.4  Filter Preparation and Analysis
2.10.5  Calculations,  Validations, and Reporting of PM10 Data
2.10.6  Maintenance
2.10.7 Auditing Procedures
2.10.8 Assessment  of Monitoring Data for Precision and Accuracy
2.10.9 Recommended Standards for Establishing Traceability

-------
2.11   Reference Method for the Determination of Particulate
       Matter as PM10 in the Atmosphere ( High-Volume PM10
       Method)

2.11.1 Procurement of Equipment and Supplies
2.11.2 Calibration Procedures
2.11.3 Field Operations
2.11.4 Filter Preparation and Analysis
2.11.5 Calculations, Validations, and Reporting of PM10 Data
2.11.6 Maintenance
2.11.7 Auditing  Procedures
2.11.8 Assessment of Monitoring Data for Precision and Accuracy
2.11.9 Recommended Standards for Establishing Traceability

-------

-------

-------
                     Addendum to Section 2.0.1
               Sampling Network Design and Site Selection
The following information should be added to Table 1.2.

     Pollutant         Spatial Scale              Characteristics
          PM
            10
 Micro
                         Middle
 Areas such as downtown street
 canyons and traffic corridors;
 generally not extending more
 than 15 meters from the roadway
 but could continue the length of
 the roadway.


 Used to evaluate possible short-
 term public health effects of
 particulate matter pollution;
 includes areas such as shopping
 center parking lots and feeder
 streets.  .
         Pb
                     Neighborhood
Micro
                        Middle
 Homogeneous urban subregion;
 dimensions of a few kilometers.

 Areas such as downtown street
 canyons and traffic corridors, and
 areas extending up to
 approximately 100 meters that
 are impacted  by plumes of
 stationary sources.

 Areas up to several city blocks
 with dimensions on the order of
 100 to 500 meters; areas may
 include schools and playgrounds
 in center city  areas that are close
to major roadways.

-------
Pollutant         Spatial Scale              Characteristics

                 Neighborhood     Homogeneous land use areas
                                  where children live and play;
                                  dimensions of 0.5 to 4
                                  kilometers.

-------
                                              Section No. 2.0.1
                                              Revision No. 1
                                              Date July 1, 1979
                                              Page 1 of 23
 1.0  SAMPLING NETWORK DESIGN AND SITE SELECTION
      Air quality samples are  generally  collected for one or more
 of the following purposes:               .         '
    .  1.    To judge  compliance -with  and/or progress made  toward
 meeting ambient air quality standards.
      2.    To activate emergency  control procedures  that  prevent
 or alleviate air pollution  episodes.
      3.    To  observe  pollution  trends   throughout  a   region,
 including nonurban  areas.
      4.    To provide  a -data  base  for research  evaluation  of
 effects;  urban,  land-use,   and transportation planning;  develop-
 ment  and evaluation  of abatement  strategies;  and development
 and validation  of diffusion models.
      With the end use of the  air  quality samples as  a  prime  con-
 sideration,  determination of one or more of the  following must  be
 basic objective(s)  of  the monitoring network:
      1.    The highest  concentrations expected in the  area covered
 by  the network.
      2.    The representative  concentrations  in areas  of  high
 population density.
      3.    The impact  of  significant sources or  source  categories
 on  ambient pollution levels.
      4.    The general background concentration levels.
 These  four objectives indicate the  nature of  the  samples  that
 the  monitoring  network will collect and that must be representa-
 tive  of the spatial area being studied.
 1-1  Monitoring Objectives  and Spatial Scales
     The  goal in siting monitors  is  to match  the  spatial scale
 represented by the sample of monitored air with the spatial scale
most  appropriate  for the   stations  monitoring objective  .(see
Tables 1.1 and  1.2).   The  representative  measurement  scales 'of
most interest for the previously stated monitoring objectives are
defined  as follows:

-------
                                             Section No. 2.0.1
                                             Revision No. 1
                                             Date July 1, 1979
                                             Page 2 of 23
     Table 1.1  RELATIONSHIPS AMONG MONITORING OBJECTIVES
                AND SCALES OF REPRESENTATIVENESS
Monitoring objective
Highest concentration
Population
Source
General/background
   Appropriate siting scales
Middle, neighborhood (sometimes

 urban) } M
Neighborhood j
Middle, neighborhood fi/cfLo
Neighborhood, regional

-------
                                              Section No^  2.0.1
                                              Revision No.  1
                                              Date  July  1,  1979
                                              Page  3  of  23
    Table  1.2
CHARACTERISITICS OF SPATIAL SCALES RELATED
        TO EACH POLLUTANT
 Pollutant
 Spatial scale
                                        Characteristics
  P1-.C
               Middle
               Neighborhood
               Regional
                     Dimensions  of a few hundred
                     meters;  e.g.,  parking lots
                     shopping centers,  stadiums,
                     office buildings

                     Homogeneous urban  subregion;
                     dimensions  of a few kilom-
                     eters; e.g.,  industrial,
                     commercial,  and residential
                     areas

                     Dimensions  of hundreds  of
                     kilometers;  sparsely populated
                     areas with  uniform surface
                     dust entrainment
SO,
Middle
               Neighborhood
               Regional
Effects of control strategies
to reduce urban concentrations
and monitoring air pollution
episodes assessed

Suburban areas surrounding
urban center or large sections
of small cities and towns;
may be associated with base-
line concentrations in areas
of projected growth

Information on background
air quality and interregional
pollutant transport
CO
Middle
                                   Dimensions of tens of meters
                                   to hundreds of meters;  e.g.,,
                                   freeway corridor or block of
                                   street development or indirect'
                                   sources (shopping centers,
                                   stadiums,  and office build-
                                   ings.)
(continued)

-------
                                             Section No. 2.0.1
                                             Revision No. 1
                                             Date July 1, 1979
                                             Page 4 of 23
Table 1.2 (continued)
Pollutant
°3 '
NO,
Spatial scale
               Neighborhood
Neighborhood
               Urban
               Regional
Middle




Neighborhood

Urban
     Characteristics
                    Homogeneous urban subregion;
                    dimensions of a few kilometers
 Information on health effect;
 information on developing, •
 testing  and revising concepts
 and models that describe
 urban/regional concentration
 patterns

 Large portions of an urban
 area; dimensions of several
 kilometers to >_50; used to
 determine trends

 Large portions of metropol-
 itan area; dimensions as much
•as hundreds of kilometers;
 used to  assess transport into
 urban area
Dimensions of hundreds of
meters to 0.5 km; character-
izes public exposure.in popu-
lated areas

Same as  for O3

Same as  for O0

-------
                                              Section No.  2.0.1
                                              Revision No.  1
                                              Date July 1,  1979
                                              Page 5 of 23

      1.    Microscale - concentrations in  air  volumes with dimen-
 sions ranging from a few meters to about '100 m.
      2.    Middle scale  - concentrations  typical  of areas up  to
 several  city blocks  with dimensions  ranging from about 100  m to
 0.5  km.
      3.    Neighborhood scale - concentrations within an extended
 area  of  the  city- that  has  relatively uniform  land uses  with
 dimensions ranging from 0.5  to 4  km.
      4.    Urban  scale  -  overall  citywide  concentrations  with
 dimensions of about 4 to  50  km;  this  scale usually requires  more
 than one  site for  definition.    -            •-
      5.    Regional  scale  -  usually 'concentrations measured  in  a
 rural area of reasonably  homogeneous  geography  that extends  from
 10's to 100's of km.
      6.    National  and global s.cales  -  concentrations  character-
 izing the  Nation and/or the  globe  as a whole.
 1 - 2   Representative' Sampling                             '.      '
      Assuring the  collection  of a  representative  air   quality
 sample depends on the  following factors:
      1.    Locating  the sampling  site,  and  determining that  the
 network size  is  consistent with the monitoring objectives.
      2.   Determining  the restraints  on  the  sampling site  that
 are  imposed by meteorology.            .'      .
      3.   Determining  the restraints  on  the  sampling site  that
 are   imposed  by  local  topography,   emission sources,  and  the
physical constraints.
     4.   Planning  sampling  schedules  that are  consistent  with
the monitoring objectives.                 ' •
1-2-1  Site  Location  -  In designing  an air  quality monitoring
program,   four criteria should be  considered either singly or in
combination  for  -site  location, depending  on the sampling objec-
tive.  Orient the monitoring sites to  measure:
     1.   The impacts  of  known pollutant  emission categories on
air quality.

-------
                                             Section No. 2.0.1
                                             Revision No. 1
                                             Date July 1, 1979
                                             Page 6 of 23

     2.   The population density relative to receptor-dose levels,
both short and long term.
     3.   The  impacts  of known pollutant  emission sources (area
and point) on air quality.
     4.   The. representative area-wide air quality.
To select locations  according  to  these criteria, i£ is-necessary
to  have  detailed  information on  the  location  of sources  of
emissions,  the  geographical  variability  of  ambient  pollutant
concentrations, the meteorological conditions,  and the population
density.  Therefore, the  selection  of the number, locations,  and
types  of sampling  stations  is a  complex  process that  defies  a
purely  objective  solution.    Furthermore,  the  variability  of
sources  and  their  intensities,  terrains,  meteorological  condi-
tions,  and  demographic  features  requires that  each network  be
developed individually.   Thus,  selection  of the network will  be
the result  of subjective judgments,  based on  available evidence
and on the experience of the decision1team.
     The sampling site  selection  process  involves considerations
of  economic,  logistic,  atmospheric,   and-  pollutant  reaction
factors  in  addition to the  motivation for and  the  objective  of
the sampling program.   None  of the factors stand alone.  Each is
dependent  in part  on  the   others.   However,   the objective  -of
the sampling  program must be  clearly defined before  the  selec-
tion  process can  be  initiated,  and  the  initial definition  of
priorities may have to be reevaluated after considerations of the
remaining  factors  and  before  the  final  .locations  are  chosen.
     Economic considerations  - The  economic  considerations  are
rather clearly defined.   The  amount  of money required  for data
gathering  (instrumentation,  installation,  maintenance,  data  re-
trieval), data analysis,  quality  assurance,  and data interpreta-
tion must be balanced against the  available monies  (current and
projected)  and  the cost-benefits  of additional  or  relocated
sampling sites.
     Logistical problems  -  The logistical  problems  involve  the
means  of obtaining,  analyzing, and interpreting the data.   Can

-------
                                             Section No. 2.0..I
                                             Revision No. 1
                                             Date July 1, 1979
                                             Page 7 of 23

the  current  staff manage  the proposed  system,  or  are more  or
fewer persons needed to accomplish the tasks?  Is the derived in-
formation necessary  on a real-time  basis,  or  can  several weeks
pass before it is available?
     Atmospheric problems -  The  atmospheric problems  pertain to
definitions  of the  spatial  and temporal  variabilities  of the
pollutants and  their transport.   Effects  of buildings, terrain,
and  heat  sources  or sinks  on the  air trajectories" can produce
local  anomalies  of   excessive   pollutant  concentrations.   Wind
velocity,   wind  shear, . and  atmospheric  stability  can  greatly
influence the dispersal of pollutants.
     Pollutant considerations -  A sampling  site or an array of
sites  for  one  pollutant may be inappropriate  for  another pol-
lutant species because of the configuration of sources, the local
meteorology,  or the terrain.  Pollutants undergo changes in their
compositions between  their  emission  and  their detection;  there-
fore, fhe impact of that change on the measuring system should be.
considered.  Atmospheric  chemical  reactions such  as the produc-
tion  of  O_ .in  the presence  of  NOv  and  HC,  -and the  time -delay
          *5                        X
between the  emission  of NO  and HC and  detection of peak  O_
                             X      •                            O
values may  require either a  sampling network  for the precursors
of O3  and/or a different network for the  actual O_  measurement.
     Summary -  While the interactions of  the' factors identified
above  are  complex,  the site  selection problems  can  be- resolved.
Experience in  the operation  of  air  quality measurement systems;
estimates   of  air • quality;   field   and  theoretical  studies  of
atmospheric  .diffusion;    and   considerations  of   atmospheric
chemistry  and  air pollution effects  are  the  requirements that
combine to  make  up  the expertise needed to  select  the optimum
sampling site for obtaining data  representative of the monitoring
objectives.
1-2.2  Meteorological Restraints  - Meteorology must be considered
in determining not only the geographical location of a monitoring
site but  also  such  factors  as  height, direction,  and extension

-------
                                              Section No. 2.0.1
                                              Revision No. 1
                                              Date July 1, 1979
                                              Page 8 of 23

 of   sampling  probes.   Meteorological  parameters   having  the
 greatest  influence  on  the  dispersion of pollutants  are  the
 direction,  speed,  and variation  of  the wind  as  well  as the at-
 mospheric stability and lapse rate.
      Wind direction indicates  the general  movement of pollutants
 in the  atmosphere.   Review of  available data can  indicate  mean
 wind direction in the vicinity of the major sources of emissions.
      The windspeed  affects (1) the  travel  time  from the  source
 to receptor and  (2) the  dilution of polluted air in the downwind
 direction.   That is,  the concentrations of  air pollutants are in-
 versely proportional to the windspeed.   Turner gives  an example:
      If a continuous  source is  emitting a certain pollutant  at
      a rate  of  10  g/s,  and the  windspeed is 1  m/s,  then in  a
      downwind length of the plume of 1 m will be contained 10  q
      of the  pollutant,  since 1  m of  air  moves  past the  sourcl
      each  second.   Next, . consider that  the conditions  of emis-
      sions  are  the  same,   but  the windspeed is  5 m/s.   In this
                                                       .
     SJ?h  nf1^  !• mi °f 'fir m°Ves past  the  source each  second,
     each  meter of plume  length contains  2  g  of pollutant.
     Wind   variability  refers   to  random  motions  in  both  the
 horizontal and vertical  velocity components  of the wind.   These
 random  motions  can be considered  atmospheric turbulence,  which
 is  either  mechanical  (caused  by structures  and  changes in ter-
 rain)  or  thermal  (caused  by heating  and cooling of land masses
 or  bodies   of  water).   if  the  scale of a  turbulent  motion is
 larger -than the size of the pollutant plume,  the turbulence will
 move  the  entire plume  and  cause  looping  or fanning.   If  the
 scale  of turbulent motion  is  smaller than the size of the pol-
 lutant  plume,  the turbulence will cause  the  plume  to diffuse or
 spread  out.   If these meteorological  phenomena impact with some
 frequency  on a  sampling site, the measured data must be evaluated
 in  light  of  possible  fumigation or  other  unusual atmospheric
conditions  .
     A useful way  of displaying wind data is a wind rose diagram
constructed to  show  the  distribution of windspeeds  and  direc-
tions.    Data  from  which  wind  rose  diagrams   can be constructed

-------
                                              Section No.  2.0.1
                                              Revision No.  1
                                              Date  July  1,  1979
                                              Page  9  of  23

 are  available in  tabulated  form  through the  National  Climatic
 Center  of  the  National Oceanographic  and  Atmospheric  Admini-
 stration  in Asheville,  North Carolina.   The wind  rose  diagram
 shown  in Figure 1.1  represents  conditions  as they converge  on
 the  center  (or site  under  consideration)  from each  direction of
 the  compass.   More complete explanations of  the construction  and
 use  of  wind   roses  are  available.-  This  brief  discussion  and
 figure  indicate  a  few meteorological effects to be  considered in'
 the  siting  of  network  surveillance  equipment.   More  detailed
 guidance  for   meteorological  considerations  is   available.1'4
 Relevant  weather  .information such  as  stability-wind  roses  are
 usually  available  from  local  National  Weather Service stations.
 In   cases  of  complex   meteorological  and   terrain situations,
 diffusion meteorologists  should be consulted.
 1-2.3  Topographical Restraints - Both the transport and-the dif-
 fusion  of  air  pollutants  are  complicated  by   topographical
 features.   Minor-  topographical   features  may exert  small   in-
 fluences; major  features,, such as  deep river valleys or mountain
 ranges, may affect large  areas.  Before, final site selection,  re-
 view  the  topography  of the area to  ensure   that  the purpose  of
 monitoring  at  that  site   will   not   be  adversely  affected.
 Table  1.3  summarizes  important  topographical  features,  their
 effects on air flow,  and some examples of influences on monitor-.
 ing site selection..
     Land use and topographical characteristics of specific areas.
 can be determined from U.S.  Geological Survey (USGS)  maps  as well
 as from land-use maps.
     Final placement  of the monitor  at  a selected  site  depends
 on physical. obstructions and  activities  in  the  immediate area;
 accessibility,  availability of utilities  and other support facil-
 ities;  and correlations  with  the  defined purpose of the specific
monitor and  the monitor design.    Because  obstructions  such   as
 trees and fences  can significantly alter  the air  flow-, monitors
should be placed  away  from such  obstructions.   It  is  important
 for  air   flow  around  the  monitor  to  be representative of the

-------
                                            Section No.' 2.0.1
                                            Revision No. 1
                                            Date  July 1, 1979
                                            Page  10 of 23
                    0-3_ 4-7  8-12 13-18  19-24
                      Speed Classes (mph)
                   0  12  3 A  5  67  8910

                           Scale; %
a,,.
 Bias removed and calms distributed.
           Figure  1.1  Wind  rose pattern.

-------
                                                       Section No.  2.0.1
                                                       Revision No.  1
                                                       Date  July  1,  1979
                                                       Page  11 of 23
       Table  1.3   RELATIONSHIPS OF  TOPOGRAPHY,  AIR FLOW, AND
                       MONITORING SITE SELECTION
Topographical
  feature
    Influence- on air flow
Influence on monitoring
    site selection
Slope/Valley
Downward air currents  at night
and on cold days;  upslope winds
on clear days when valley
heating occurs

Slope winds and valley
channelled winds;  tendency
toward down-slope  and  down-
valley winds; tendency
toward inversions
Slopes and valleys  as  special
sites for air monitors
because pollutants  generally
well dispersed;  concentration
levels not representative  of
other geographic areas;  pos-
sible placement  of  monitor to  '
determine concentration  levels
in a population  or  industrial
center in a valley
Water
Sea or lake breezes  inland
or parallel to shoreline
during the day or in cold
weather;, land breezes  at
night          .   ,       '
Monitors on shorelines  gener-
ally useful for background
readings or for obtaining
pollution data on water
traffic
Hill'
Sharp ridges causing  tur-
bulence;  air flow around
obstructions during stable
conditions,  but over  obstruc-
tions during unstable condi-
tions
Depends on source orientation;
upwind source emissions  gener-
ally mixed down the  slope,
and siting at foot of  hill  not
generally advantageous;  down-
wind source emissions  generally
downwashed near the  source;
monitoring close to  a  source
generally desirable  if popula-
tion centers adjacent  or if
monitoring protects  workers
Natural or
 manmade
 obstruction
Eddy effects
Placement near obstructions
not generally representative
in readings

-------
                                              Section No.  2.0.1
                                              Revision No.  1
                                              Date July 1,  1979
                                              Page 12 of 23

 general air flow in the  area  to prevent sampling bias.   Detailed
 information on urban-physiography (e.g.,  buildings,  street dimen-
 sions)   can be  determined  through visual  observations,  aerial
 phst-ography,  and surveys.   Such information can be  important in
 determining the  exact  locations  of  pollutant  sources  in  and
 around  the  prospective monitoring site  areas.
      Network designers  should  avoid  sampling locations that  are
 unduly  influenced by downwash or by  ground dust  (e.g.,  a  rooftop
 "air inlet near a stack or  a  ground-level inlet near an  unpaved
 road);  in these  cases,  either elevate  the  sampler intake  above
 the level  of  the maximum  ground turbulence effect or pl-ace  it
 reasonably  far  from  the source  of ground  dust.
      Depending  on the  defined  objective—that  is,  to determine
 the background  levels, or to determine  the maximum concentrations
 and so  forth—the monitors .(for sampling at  a  particular  site)
 would be placed  according to  exposure  to  pollution,  including
 exposure  to motor  vehicle emissions.   Therefore in " most  every
 instance, a practical consideration  must be  made of unavoidable
 physical  restraints  on the  optimum representativeness of  sample
 collection.   This  consideration should  include categorization  of
 sites relative  to their  local  placements.  Suggested, categories
 relating to sampling site placement for measuring a corresponding
 pollutant impact are as follows:
     Category A (ground level station)  -  Heavy, pollutant concen-
 trations—high  potential  for pollutant buildup.   A site that  is
 3 to  5  m>(10-16 ft)  from major traffic artery and that has  local
 terrain  features  restricting ventilation.  A  sampler probe that
 is 3 to 6 m (10-20 ft) above ground.
     Category B (ground level station)  -  Heavy pollutant concen-
trations—minimal potential for a pollutant buildup.  A site 3  to
15  m  (15-50 ft)  from a major  traffic  artery with  good natural
ventilation.  A sampler probe  that  is 3 to 6 m  (10-20  ft) above
ground.

-------
                                              Section No. 2.0.1
                                              Revision No. 1
                                              Date July 1, 1979
                                              Page 13 of 23

      Category C (ground level station) -  Moderate pollutant con-
 centrations.   A site that  is 15 to 60 to (50-200 ft) from a major
 traffic artery.   A sampler  probe  that  is 3  to 6 m  (10-20 ft)
 above ground.
      Category D (ground level station)  -   Low  pollutant  concen-
 trations.   A site that  is  >_60 m (>200 ft) from a traffic artery.
 A sampler  probe that is 3 to 6 m (10-20 ft) above ground.
      Category E (air mass station)  -  Sampler probe  that is  be-
 tween 6 and 45 m  (20  and 150 ft) above ground.   Two  subclasses:
 (1)  good exposure  from all  sides  (e.g.,  on top  of building)  or
 (2)  directionally biased  exposure  (probe  extended  from window).
      Category  F (source-oriented station)   -  A   sampler  that  is
 adjacent  to   a  point  source. •  Monitoring   that yields   data
 directly .relatable  to emission source.
 1-2-4  Sampling Schedules -r  Current Federal regulations' specify
 the  frequency  of sampling  for. criteria pollutants to meet minimum
 SIP  (State  Implementation Plan) surveillance requirements .'   Con-
 ^MU°/^  /Lamp}^n1-^s  sPecified  except for 24-h ' measurements  of
 ,"' «  tOj rb OL.MD Tor%
 total  tfLionondod fm-t-i ml -it- nr (Tgp)  -.1-.^ a/|-h  integrated  valuoc  of.
 SO^  and NO^ .    n-n^  K-J »~-\  nri^  ^nc  impingar maacurcmcntc  arc ro
 quirod  at — leact — once — every  6  dayc - -an • equivalent of  about  61-
 random  oamploo/yE.   The 24-h samples should be  taken from  mid-
 night  (local standard time) to midnight and thus  shou-ld  represent
 calendar days  to permit the direct use  of the sampling data  in
 standard daily  meteorological summaries.   The following  are  rec-
 ommended frequencies  for noncontinuous hi-vol  and impinger  sam-
plings to adequately define TSP, SO2 ,  and NO2 levels:
     1.   The  most polluted  . sites  in an  -urban  area  should be.
sampled  at frequencies  greater  than the  minimum requirements."
     2.   Sites where  the highest 24-h and  annual  averages are
expected should' yield the most frequent
     3.   Areas of  maximum SO2  and  N02  concentrations should be
sampled using  continuous  monitors in  place  of SO2/NO2 impingers
if possible.

-------
                                              Section No.  2.0.1
                                              Revision No.  1
                                              Date  July  1,  1979
                                              Page  14 of 23

     4.   Noncritical  sites  (i.e.,  sites other than maximum  con-
centration  sites)  can be sampled  intermittently.  Intermittent
sampling  calls for  adopting  a systematic sampling schedule  that
considers  statistical  relationships  for  characterizing  an  air
pollutant  for  a  given  time  period and area. '    Actually any
schedule  which gives  61  samples/yr and 5/quarter (in  accordance
with item  6  below)  is satisfactory, but not  as convenient -as the
systematic schedule  of every  6th day,  for example.
     5.   Downwind  sites monitoring  for SO_, NO2/  and *S3»Afrom
isolated point  sources should use continuous instruments for the
gaseous pollutants,  and should sample at least once every 6 days
for •£§£-. /vj-AT/cuL/f-rs:  M/hTT££.
     6.   The minimum numbers of samples required for  the appro-
priate  summary  statistics should be  taken.   At  least 75% of the
total  possible  observations  must be  present before summary sta-
tistics are calculated.  The  exact requirements follow:7
     Time interval        Minimum number of observations/averages

3-h running average           3 consecutive hourly observations
8-h running average           6 hourly observations
24 h                        18 hourly observations
Monthly .                   21 daily averages
Quarterly                     3 consecutive monthly .averages
Yearly                    •    9 monthly averages with at least
                               2 monthly averages/quarter

For intermittent  sampling data,  there must be at least five ob-
servations/quarter;  if one month  has  no  observations,  the re-
maining two months must have  at least two.
     7.   If validation procedures  indicate  that the criteria in
item 6  are  not  fulfilled (the minimum numbers must  be valid ob-
servations), the  sampling  frequency  should  be  increased during
the period in which corrective measures are being pursued.

-------
                                             Section No. 2.0.1
                                             Revision No.  1
                                             Date July 1,  1979
                                             Page 15 of 23

     More  extensive  treatments  of   sampling  frequencies,  as
related to data analysis,'are in References 6, 7, and 8.
     Table  1.4  lists approximate  numbers  of stations for each
NAMS area,  as determined by population and concentration catego-
                   P/M-/O                                        9
ries,  for  SO2  and ¥Si>  as specified in  Appendix D, 40 CFR 58.
1.3  Sampling Site and Equipment Requirements
     Sampling  site  and  equipment  requirements  are  generally
divided  into  three  categories,  consistent  with  the  desired
averaging- times:
     1.   Continuous—Pollutant  concentrations  determined  with
automated methods, and recorded or displayed .continuously.
     2.   Integrated—Pollutant  concentrations  determined  with
manual  or  automated methods   from  integrated  hourly or   daily
samples on a fixed schedule.
     3. •  Static—Pollutant  estimates  or   effects   determined
from   long-term   (weekly  or  monthly)  exposure  of  qualitative
    -m—
measurement devices or materials.
     Air  quality monitoring sites  that use  automated equipment
to  continually   sample  and  analyze  pollutant  levels  may  be
classified as primary.  Primary monitoring stations are generally
located  in  areas where pollutant  concentrations are  expected
to  be  among the  highest and in areas with the highest population
densities;  thus,  they are often used  in health effects research
networks;  in addition,  these stations are  designed as a part of
the air pollution episode warning system.
1.4  Quality Assurance
     The .quality assurance  plan*  should  include  specific  docu-
mentation  of site  characteristics  for  each  monitoring station.
*Minimum  suggested content for  such a plan  is discussed in the
 Quality Assurance Handbook for Air Pollution Measurement
 Systems,  Vol.  I,  Section  1.4.23,  entitled  "Quality Assurance
 Plans  for Projects and Programs."

-------

^^^v
"5s
a.

£3
O
'.p
(0
u.
c
CD
O
C
o
O
o
•wj
o
C
o
*4->
03
v_
•M
C
0)
o
c
o
o
E
3
•a
CD
^c
03
+->
C
CD
O
C
o
U
JC
O)
i


o
CO
fl^
\Lf
03
O
c
g
_£0
13
CL
O
a.


0
^
CL




ts
0
CO




o
-.-.,
^^>
DL





^
O
CO
o
"5
Q.



CO
4— '
c
CD
O
C
o
CJ
JC
CO
IE
NAAQS; PM10
>.
CO
c
o
o
CD
CO
'o

c^
O
O
*""
O
CO
CJ
<
z
CO
a.
"o
O
CD
C
"T3
CD
CD
O
X
CD
01
O
CO •
CO
cO
o <.
co <-
CD a""
o O
§ro
o co

.=!. CD
"O 03
CD 0
^ CD
C
ca
CO
CO
0
CL
. „
CO
a
*^L
*^L
z
>.
. CO
•a
c
o
o
CD
CO
H-
o
•s^
O
o
v
^«
o
CO
a
^f ,
<^r
z
03
E
'u.
Q.
14_
O
0
CD
V

Q
CO
c
'•p
03 .
4= CO

o "^
0 <
o ^
> ^-
0 0
—I CD

-------
                                             Section No. 2.0.1
                                             Revision No. I
                                             Date July 1, 1979
                                             Page 17 of 23

This  information will assist  in providing objective inputs into
the  evaluation of data  generated at that  site.   Typically, the
                                          2
site  identification record should include:
      1.   Data   acquisition  objective:   air  quality  standards
monitoring.
      2.   Station type (stationary, mobile, etc.);
      3.   Instrumentation  checklist  ("manufacturer's model  num-
ber,  pollutant measurement technique, etc.).
      4.   Sampling system (probe type, height, flow, etc.).
      5.   Spatial  scale  of  the  station  (site  category—i.e.,
urban/industrial, suburban/commercial,  etc.;  physical location-.-
i.e., address, AQCR, UTM coordinates, etc.).
      6.   Influential pollutant  sources  (point and area sources,
proximity, pollutant density, etc.).
      7.   Topography  (hills, valleys,  bodies  of  water,  trees';
type  and  size,  proximity, orientation,  etc.;  picture of  a  360°
view  from the probe of the monitoring site).
      8.   Atmospheric   exposure    (unrestricted,   interferences,
etc.).
      9.   Site diagram  (sample  flowsheet,  service  lines,  equip-
ment  configuration,  etc.).
1.5  Network Designs Examples
     As previously noted, networks  are  designed for at least one
of  four major purposes  (Subsection  1.0).   The following tabula-
tion presents examples of currently implemented networks applica-
ble to each of these four.

-------
                                             Section No. 2.0.1
                                             Revision No. 1
                                             Date July 1, 1979
                                             Page 18 of 23


Monitoring
objective                Network                  Comment

Compliance          SIP  (State  Imple-   To demonstrate attainment
                    mentation Plan)     or maintenance of AQS

Emergency           SIP/local, agency    To activate immediate,
 episode            emergency control   short-term emission
                    program'            controls for episode
                                        prevention

Trend    .           NASN (National      To fulfill mandate of
                    Air Sampling        Federal legislation
                    Networks)

Research           , CHAMP  (Community    To determine long-term
                    Health Air Monitor- pollutant trend in
                    ing Program)        selected areas with
                                        health effects

1•5•1   Compliance Monitoring  -  The  information  required  for
selecting  the  number  of  samplers  and  the  sampler  location is
essentially  the  same—that is,  isopleth  maps,  the  population
density maps,  and the source locations.   Following  are  the sug-
gested guidelines:

     1.   The  priority  area is  the  zone  of  highest  pollutant
concentration  within  the  region;  one or more stations  are to
be located in this area.

    •2.   Close  attention  should be given to  densely  populated
areas within the region, especially when they are in the vicinity
of heavy pollution.

     3.   The  quality, of  air  entering the  region  is to  be as-
sessed  by  stations  situated  on the  periphery  of  the  region;
meteorological factors (e.g., frequencies of wind directions) are
of primary importance in locating these stations.

     4.   Sampling  should  be  undertaken  in  areas of  projected
growth  to  determine  the  effects  of  future  development  on the
environment.

     5.   A  major  objective of  surveillance  is  evaluation  of
progress made   in  attaining the  desired  air  quality;   for  this
purpose, sampling stations  should  be strategically  situated to
facilitate  evaluation of the implemented control tactics.

-------
                                              Section No.  2.0.1
                                              Revision No.  1
                                              Date July 1,  1979
                                              Page 19 of 23

      6.    Some information of air quality  should be available to
 represent all portions of the regions.
 Some stations will be  capable of fulfilling more than one of the
 functions indicated;  for example, a  station located in a  densely
 populated area can  indicate  population exposures and'also docu-
 ment  the changes  in  pollutant  concentrations  resulting  from
 control  strategies  used in the area.
 1-5.2   Emergency Episode Monitoring  - For episode avoidance  pur-
 poses,  data  are  needed quickly--in no less  than  a few hours after
 the  sensor is contacted  by the  pollutant.   While it  is possible
 to  obtain . data  rapidly  by  on-site manual data  reduction  and
 by  telephone reporting,  there is a  trend toward using automated
 monitoring  networks.   The  severity  of  the  problem,  the  size  of
 the  receptor area, and the .availability of resources all  influ-
 ence both the scope and  sophistication of  the monitoring  system.
     It  is  necessary  to  use  continuous air samplers because  an
 episode  lasts only  a  few  days  and  because the control  actions
 taken must be based on  real-time measurements that .are correlated
 with the decision criteria.  Based on alert criteria now  in  use,
 1-h averaging times are adequate  for  surveillance of episode  con-
 ditions.   Shorter averaging  times  provide  information on data-
 collecting excursions,  but they  increase the need for automation
 because  of the bulk of  the  data obtained.  Longer averaging times
 (>6  h)  are not  desirable because of the delay  in response  that
 these  impose.  After   an  alert   is   announced,   data  are   needed
 quickly  so  that  requests  for information   on the event  can be
 responded to.                                              ,
     Collection and analysis  must be accomplished rapidly  if the
 data are to be useful immediately.  There is no time to check out
 the methods,  to  run  blanks, to. calibrate and so forth, after the
 onset of episode conditions.  For the instrument to be maintained
 in peak  operating condition,  either  personnel must  be stationed
 at the  sites  during the  episode  or  automated equipment  must be
operated  that  can  provide  automatic  data- transmission  to  a
central location.

-------
                                              Section No.  2.0.1
                                              Revision No.  1
                                              Date  July  1,  1979
                                              Page  20 of 23

      The instruments employed for measuring  the pollutant- param-
 eter use  either  wet chemical  techniques or  physical  methods.
 Chemical analysis  always  involves  the • use  of  consumable  sup-
 plies.   The  chemicals must be  replaced  on  a schedule that is  con-
 sistent  with their  stability  and  with  the rate at  which  samples
 are  taken.   Currently used instruments require adequate  supplies
 of chemicals  for  operation. for  3 mo  so that  the supplier can
 comply >  with  delivery  schedules.    In some cases,   analytical
 reagents for  specific air contaminants  deteriorate rapidly and
 should  have. protective  storage.   Physical methods  are performed
 with relatively complex equipment that  must be  installed  cor-
 rectly  and" cared  for by trained  personnel;  the  accuracy of the
 equipment  is  affected by  mechanical  shock,  ambient temperature
 extremes,  voltage  supply  stability,  dirty or dusty atmospheres,
 and  corrosive  chemicals.
      Episode conditions  threaten human  health and welfare.  Moni-
 toring  sites should  be  located in areas  where  human health and
 welfare  are most threatened:
      1.   In densely populated  areas.
      2.   Near large stationary sources of pollutants.
      3.   Near hospitals'.
      4.   Near high-density traffic interchanges.
      5.   Near homes  for the aged.
 A  network  of  sites  is  useful  in  determining   the  'range  of
 pollutant concentrations within an area.  Although the most de-
 sirable monitoring sites are not necessarily  the most convenient,
 consideration  should be  given for reasons of  access,  security,
 and  existing  communications  'to   the  use  of public  buildings:
 schools,  firehouses,  police  stations,  hospitals,   and water or
 sewage plants.
 1-5.3   Trend monitoring  -  As  typified  by  the  National  Air
 Surveillance Network  (NASN),   trend  monitoring is  characterized
by locating  a  minimal number  of monitoring sites across as large
 an area   as possible.   The  program  objective  is  not  only to

-------
                                              Section No.  2.0.1
                                              Revision No.  1
                                              Date  July 1,  1979
                                              Page  21 of 23

 determine,  in a broad sense  the  extent and nature  of  air pollu-
 tion  but also to determine the variations  in the  measured levels
 of  atmospheric  contaminants  in  respect  to geographical, socio-
 economic,  climatological,  and other  factors.  The.data  acquired
 are useful  in planning epidemiological  investigations and in pro-
 viding  the  background, against which more intensive  community and
'statewide studies of  air pollution can  be-conducted.
      Urban  sampling  stations are  usually  located  in the  most
 densely populated  areas  of a region.   In most regions, there are
 several urban sites.
      Nonurban station  locations  include  various  topographical
 categories  such " as  farmland,   desert,  forest,  mountain,  and
 coast.   Nonurban stations  are not  specifically  selected  to be
 "clean  air" control  sites  for urban areas,  but they  do provide.
 a' relative comparison  between  some urban  and  nearby  nonurban
 areas.
      in interpreting  trend data,  one must- consider the limita-
 tions" imposed by the  network design.  Even  though precautions are
 taken, to ensure that each sampling  site  is as  representative as
possible of the designated area, it  is  impossible to be totally
certain that  the measurements obtained  at a specific site are
not sometimes unduly  influenced  by -local  factors.  Such factors
might include topography, structures, and sources of pollution in
the  immediate vicinity  of  the   site,  and  other  variables—the
effects  of  which cannot always  be  accurately  anticipated,  but
should  be considered  in  network  design.  It must be kept in mind
that  comparisons among  pollution' levels for various  areas  are
valid only if the sites are comparable.
1.-5.4   Research  Monitoring  -  An  example of  a  research-oriented
air  quality  monitoring  effort   is  EPA's  Community  Health  Air
Monitoring  Program  (CHAMP),  which  has  provided data to  develop
criteria for  both  short-  and long-term  air quality  standards.
Air monitoring  networks  related  to  health effects  are composed

-------
                                              Section No.  2.0.1
                                              Revision No.  1
                                              Date  July 1,  1979
                                              Page  22 of 23

 of integrating  samplers  both  for  determining pollutant  concen-
 trations  for >24 h  and  for developing long-term  (>_24  h)  ambient
 air  quality  standards.   This  research  requires that  monitoring
 points  be located so  that  the  resulting data will  represent the
 population  group under  study;  thus the  monitoring stations, are
 established   in  the  centers of  small  well-defined  residential
 areas  within a  community.   Data  correlations  are  made   between
 observed  health  effects and observed air  quality exposures.  •
     Requirements for  aerometric  monitoring in support of health
 studies are:
     1.   Station must be located in or near  the population under
 study.
     2.   Pollutant sampling averaging times must be sufficiently
 short to  allow  for  use in  acute health effects studies that form
 the scientific basis for short-term standards.
   .  3.   Sampling frequency, usually daily, should be sufficient
 to characterize air quality as a function of time.
     4.   System should  be  flexible and  responsive to emergency
conditions with data available on short notice.

-------
                                             Section No.  2.0.1
                                             Revision No.  1
                                             Date July 1,  1979
                                             Page 23 of 23
1.6  References
     1.    Guidelines of Air Quality Monitoring Network  Design and
          Instrument   Siting.    U.S.   Environmental   Protection
          Agency,  Office  of Air Quality Planning and  Standards.
          OAQPS No.  1.2-012.   Revised.   September 1975.   Draft.

     2.    Air  Quality  Monitoring  Site  Description   Guideline,
          Environmental  Protection  Agency,   Research  Triangle
         .-Park,  N.C.  OAQPS No.  1.2-019,  1974.   Draft.

     3.    Code  of  Federal Regulations  40.    Protection  of  the
          Environment.  Parts  50 to 69.   Revised  July 1,  1975.

     4.    Turner,   D.  B.    Workbook  of  Atmospheric   Dispersion.
          Estimates-.   Environmental   Protection  Agency.    7th
          printing,  January 1974.

     5.    Ludwig,  F.  L.  and J.  H.  S.  Kealoha.  Selecting  Sites
          for  Carbon  Monoxide   Monitoring.    EPA-450/3-75-077.
          September 1975.

     6,..    Hunt,  W. F.  The Precision Associated with the  Sampling
          .Frequency  of Log  Normally  Distributed  Air  Pollutant
          Measurements.  JAPCA 22.   September  1972.

     7.    Guidelines  for   the  Evaluation  of  Air  Quality  Data.
          Environmental Protection Agency,  Office of Air  Quality
          Planning  and Standards.   OAQPS  No.  1.2-015.   January
          1974.   p.  21.

     8.    Akland,  G.   Design  of Sampling  Schedule.   JAPCA  22.
          April  1972.              .                           —

     9.    40  CFR 58, Appendices  C -  E.


    Guideline for Lead Monitoring in the Vicinity of Point Sources  EPA-
    45 0/4-81-00 6.

    .Optimum Sampling Site Exposure Criteria for Lead. EPA  450/4-84-
    012.  February 1984.

    Network Design and Optimum Site Exposure Criteria for  Particulate
    Matter.  EPA 450/4-87-009.  May 1987.

-------

-------
                          Addendum to Section 2.0.2
••                         Sampling Considerations
    The following information should be added to table 2-2:

Pollutant


Pb






*



r.
i
• '








Scale


Micro










Middle,
neighbor-
urban,
and
regional







Height above
ground,
meters

2-7










2-15









Distance from supporting structure,
meters

Vertical
-










-










Horizontal"
>2










>2









=======
Other spacing criteria


1. Should be >20 meters from
trees.
2. Distance from sampler to
obstacle, such as a building, most
be at least twice the height that the
obstacle protrudes above the
sampler.
3. Must have unrestricted airflow
270° around the sampler.
4. No furnace or incineration fhies
should be nearby. b
1. Should be >20 meters from
trees.
2. Distance from sampler to
obstacle, such as a building, must
be at least twice the height that the
obstacle protrudes above the
sampler.
3. Must have unrestricted airflow
270° around 'the sampler.
4. No furnace or incineration flues
should be nearby."
5. Spacing from roads varies with
traffic (see Table 4 of Appendix

-------

Pollutant


PM!0












t




.


PMW
















Scale


Micro




















Middle,
neighbor-
hood.
urban,
and
regional
scale.









Height above
ground,
meters

2-7




















2-15















Distance from supporting structure,
meters

Vertical
.




















.
















Horizontal*
>2




















>2















I
Other spacing criteria


1. Should be >20 meters from the
dripline and must be 10 meters
from the dripline when the tree(s)
acts as an obstruction.
2. Distance from sampler to
obstacle, such as buildings, must
be at least twice the height the
obstacle protrudes above the
sampler except for street canyon
sites'*.
3. Must have unrestricted airflow
270° around the sampler except for
street canyon sites.
4. No furnace or incineration flues
should be nearby.
5. Spacing from roads varies with
traffic (see Figure 2 of Appendix
E°) except for street canyon sites
which must be from 2 to 10 meters
from the edge of the nearest traffic
lane.
	 — — 	 . 	
1. Should be >20 meters from the
. dripline and must be 10 meters
from the dripline when the tree(s)
act as an obstruction.
2. Distance from sampler to ;
obstacle, such as buildings, must
be at least twice the height the
obstacle protrudes above the
sampler1".
3. Must have unrestricted airflow
270° around the sampler.
4. No furnace or incineration flues
should be nearfayb.
5. Spacing from roads varies with
traffic (see Figure 2 of Appendix
E"). ' 1

-------
   Pollutant
  Scale
Height above
  ground,
   meters
Distance from supporting structure,
            meters
                                            Vertical
                                           Horizontal*
                                                                                Other spacing criteria
VOC
PAMS
                               3-15
                                              >1
                                                              >1
                                                          1. Should be' >20 meters from the
                                                          dripline and must be 10 meters
                                                          from the dripline when the tree(s)
                                                          act as an obstruction.
                                                          2. Distance from probe inlet to
                                                          obstacle must be at least twice the
                                                          height the obstacle protrudes above
                                                          the inlet probe.
                                                          3. Must have unrestricted air flow
                                                          in an arc of at least 270° around
                                                          the probe inlet and the predominant
                                                          wind direction for the period of
                                                          greatest pollutant concentration (as
                                                          described for each site in section
                                                         4.2 of appendix D*) must be
                                                         included in the 270° arc.  If probe
                                                         located on the side of a building
                                                         unrestricted air flow must be 180°.
                                                         4.  Spacing from roadways varies
                                                         with traffic (see Table 5 of
                                                         Appendix E*).
     When a probe is located  on a rooftop,  this separation is  in reference to walls, parapets  or
     penthouses located on the roof.
     Distance is dependent on the height of the furnace or the incineration flue, the type of fuel
     or waste burned, and the quality of the fuel  (sulfur, ash  or lead content).   This is to avoid
     undue influences from minor pollutant sources
     40CFR58.
     Sites  not meeting this criterion would be classified as middle scale.

-------

-------
                                             Section No. 2.0.2
                                             Revision No. 1
                                             Date July 1, 1979
                                             Page 1 of 9
2.0  SAMPLING CONSIDERATIONS

2.1  Environment Control

     A proper sampling  environment  demands  control of all physi-

cal parameters  external to the samples that  might affect Sample

stability, chemical reactions within the sampler,  or the function

of sampler components.  The important parameters to be controlled

are summarized in Table 2.1.
          Table 2.1  ENVIRONMENT CONTROL PARAMETERS
Parameter
   Source of
 specification
  Method of
    control
Instrument vibration
Manufacturer's
  specifications
Design of instru-
 ment housings,
 benches,  etc.  ,
 per mfr.  spec.
Light
 Method description
  or manufacturer's
 • specifications
Shield chemicals
 or instruments
 that can be
 affected by nat-
 ural or artifi-
 cial light
Electrical voltage
Method description
 or manufacturer's
 specifications
Constant voltage
 transformers or
 regulators; sep'a-•
 rate power lines;
 isolated high cur-
 rent drain eguip-
 ment such as hi-
 vols, heating
 baths, pumps from
 regulated circuits
Temperature
 Method description
  or analyzer
  specifications
Regulated air con-
 ditioning system;
 24-hour temperature
 recorder; use
 electric heating
 and cooling only
 (see note on next
 page)
Humidity
 Method description
  or analyzer
  specifications.
Regulated air con-
 ditioning system;
 24-hour recorder

-------
                                              Section No.  2.0.2
                                              Revision No.  1
                                              Date July 1,  1979
                                              Page 2  of 9

 Note;   With respect  to environmental temperature for designated
 analyzers,  most  such analyzers  have been  tested and qualified
 over a temperature range  of  20°C (68°F) to 30°C  (86°F).   (A few
 analyzers  are  qualified over a  wider range.)  This  temperature
 range  specifies both  the  range of acceptable  operating  tempera-
 tures  and  the range of  temperature change  which the  analyzer can
 accommodate without excessive drift.   It is the latter—the range
 of temperature change  that may occur between zero  and span ad-
 justments—that is, by  far,  the most important.  To  accommodate
 current  energy conservation regulations or  guidelines  specifying
 lower  thermostat  settings, designated analyzers located  in faci-
 lities subject to these restrictions may be operated  at  tempera-
 tures  down to  18°C  (64°F),  provided the  analyzer temperature does
 no.t  fluctuate  by  more than 10°C  (e.g.,  18°  to  28°C) between zero
 and  span  adjustments.   However,'  operators  should  be  alert to
 situations  -where  environmental  temperatures  might  fall  below
 18°C  (64°F),  such  as  during  night hours  or weekends.  Use of a
 thermograph at the  analyzer site may be necessary to  detect such
 situations.  Temperatures  below 18°C may necessitate supplemental
 temperature  control  equipment or  rejection of the  area  as  an
 analyzer installation site.
 2-2  Sampling Probes and Manifolds
 2-2-1  Design of Probes and Manifolds1 - Some important variables
 affecting the  sampling  manifold design are  the  diameter',  length,
 flow  rate,  pressure  drop,  and materials of construction.   Con-
 siderations for these parameters are  discussed below for both a
.vertical laminar flow and  a conventional manifold design.
     Vertical laminar flow design - By  the  proper selection of a
 large diameter  vertical  inlet probe and by maintaining a laminar
 flow throughout,  the  sample  air  is  not permitted  to react with
 the walls  of  the probe.   Numerous  materials  such  as glass, PVC
plastic,   galvanized  steel,   and  stainless  steel,  can be used
 for constructing  the  probe.   Removable  sample lines constructed
of Teflon or glass can be used to provide each device with sample

-------
                                             Section No.  2.0.2
                                             Revision No.  1
                                             Date July  1,  1979
                                             Page 3 of  9

 air.   A flow rate of 5 3,/min  (0.18  ft3/min)  in 1.5-cm (0.6 in.)
 diameter  tubing  (commonly  used in  monitoring stations)  is not
 satisfactory  for  this application because of almost complete dif-
 fusion losses.
     Diameters  from 1.5  to 2.5 cm  (0.6  to  1.0 in.) with 50 to
 150. £/min  (1.8  to 5.4  ft /min)  are  unacceptable  because of high
 pressure drops.   Therefore, inlet line diameters of 15  cm  (6 in.)
 with  a flow  rate of  150 £/min (5.4  ft3/min) are  necessary if
 diffusion  losses  and  pressure drops  are to  be  minimized.   The
 sampling rate should be  maintained  to insure  laminar  flow con-
 ditions.  •           .               :        •
     Figure  2.1  is  an  example of a vertical  laminar  flow mani-
 fold.  This configuration has  the following advantages:         '  '
     1-   A  15-cm (6 in.) pipe can  be  cleaned easily by pulling
 a cloth through it with a string.
    . 2".   Sampling ports  can be cut into- the  pipe  at  any loca-
 tion  and, . if unused,  can  be  plugged  with  .stoppers of  similar
 compositions.
     3.   Metal poses no breakage hazard.
     4.   The pipe does not have to be clean to provide a repre-
 sentative sample,  as in the case with smaller tubes.
     Conventional manifold design - In practice, it may be diffi-
 cult  to achieve  .vertical  laminar   flow  because  of the- elbows
within  the   intake  manifold system.   Therefore,   a  conventional
manifold system should  be constructed  of inert materials such as
Pyrex  glass   and/or  Teflon, and  in modular  sections  to enable
 frequent cleaning.   The system (Figure  2.2)  consists of a verti-
cal "candy cane"  protruding through  the roof of the  shelter with
a horizontal  sampling manifold connected by a tee to  the vertical
section.  Connected  to  the  other  vertical outlet of  the tee is a
bottle  for  collecting heavy particles  and moisture before  they
enter  the  horizontal  section.   A  small   blower,   1700 £/min
 (60 ft /min)   at   0 cm  of water  at  static pressure,  is   at  the
exhaust end  of  the  system  to  provide  a  flow  through the system

-------
                                           Section No.  2.0.2
                                           Revision  No.  1
                                           Date July 1,  1979
                                           Page 4 of 9
 1 - 2 m
(3-6 ft)
                        15 cm
                        (6 In.)
                                     ROOF
                                       SAMPLE PROBES
                                    INCLINED MANOMETER
                                        ORIFICE METER
                    BLOWER - 150 fc/min    FOR FLOW MEASUREMENT
  Figure 2.1   Vertical laminar flow  manifold.

-------
                                              Section No. 2.0.2
                                              Revision No. 1
                                              Date July 1, 1979
                                              Page 5 of 9
  BLOWER-^
             MODULAR SECTION
MOISTURE TRAP ->
          Figure  2.2   Conventional  manifold system.

of  approximately  85   to  1401/min  (3  to  5ft3/min).   Particu-
late  monitoring  instruments,   such  'as  nephelometers,   should
each  have - separate,  intake  probes  that  are   as  short  and  as
straight as possible to avoid particulate losses due  to  impaction
on the walls of the probe.
     A recent investigation2 has shown that  there are no signifi-
cant  losses  of reactive  gas  (O3)  concentrations in.  conventional
13  mm  (0.5  in.)  inside  diameter  sampling lines   of  glass  or
Teflon, if  the sample  residence time  is 10  s  or less.  This  is
true  even  in sample lines  up to 38  m (125 ft) in length, which
collect  substantial  amounts  of  visible  contamination  due  to
ambient aerosols.   However, when  the  sample residence  time ex-
ceeds 20  s,  loss is detectable, and  at  60  s  the loss is nearly

-------
                                              Section  No.  2.0.2
                                              Revision No.  1
                                              Date  July  1,  1979
                                              Page  6 of  9

 complete.   Conversely,  when the  sample  lines  are free of  contami-
 nation,  there is no detectable  loss  at sample  residence  times up
 to  120  s.
 2.2.2   Placement of Probes and Manifolds  -  Probes and manifolds
 must be placed to  avoid  introducing bias to the  sample.  Impor-
 tant considerations  are   probe  height  above  the ground, probe
 length  (for horizontal probes),  and physical influences  near the
 probe.   Some general guidelines for probe and  manifold placement
 are:
     1.   Probes  should not be placed next to air  outlets  such as.
 exhaust fan openings          "                       •
     2.   Horizontal  probes  must extend  beyond  building over-
 hangs .
     3.   Probes  should  not be near physical  obstructions  such
 as  chimneys which can affect the air flow in the  vicinity of the
 probe.
     4.   Height  of the  probe   above  the ground  depends on the,
 pollutant being measured.                     •
     Table  2.2  summarizes'probe siting criteria specified in Ap-
 pendix  E, 40'CFR  58 for NAMS  and SLAMS.5
                3
 2.3  Maintenance
     After  an adequately  designed sampling probe  and/or manifold
 has  been selected and  installed,  the  following steps  will  help
 in maintaining constant sampling conditions:
     1.   Conduct  a leak  test  - For the  conventional  manifold,
 seal all ports and pump  down to approximately  1.25 cm (0.5 in.)
 water gauge vacuum,  as indicated by  a  vacuum gauge or manometer
 connected to  one  port.   Isolate the system.  The vacuum measure-
 ment should show no change  at the end of a 15-min period.
     2.   Establish cleaning  techniques  and  a schedule - a large
 diameter manifold may  be  cleaned by pulling  a  cloth  on a string
 through  it.   Otherwise  the manifold  must be disassembled period-
 ically  and  cleaned  with soap and water.  Visible dirt should not
be allowed to accumulate.

-------
Section No. 2.0.2
Revision No. 1
Date July 1, 1979
Page 1 of 9


ITERIA
tf
O
O
M
E-i
M
01
pa
§
o<
r%.
O




Sj
D
CO



.n

EH
















'•• r


O

e «
U ,J
o e
Q. O
CU tt
3 •-•
M IM
.2
£
0 U
c — «
«• w
^ >
41
•i i


c
-« 3
41 O
as *•
•o
41
U


C
3
O
a,

LU
X
M — •
o .e -o
T3 U *j c
.• £ * •; s.
M W CLf C.
^o e • w •<
T3 C. (A J3 — *—
•- U t_ o
3 — j£ v > —
M w -^ 41 — U
« « G J3 j= 3
J= 3 |- M
jc c o -a •—
0 l, — U.
30 «a 3 - •«— '
- -C 0 M
-> • JC -0 4)
O M «C — •
4» •* r- o i*
• — — • c* in i* u
OS* 41 M
<• — i 3 E
W fc 0 — 0 —
Ob U C • J
0 — * — G C.
**• « ^ ^ C W
Q — w u C. C
— *•= U C M 4
E"* C T "£ E *-
 * > 0
o o -• «i c «r
— « « ~ 41 3 "~ jC
3 w w > w w w M
O M M O M M — .
JC — 3 J= 3' O 3 t)
w c .e czz=:.=



o*
A

"


I .


JQ
i

(M




_
•<



a.
to


u
J3
MM O
M U U
c -a a.
— 3
•o u *j
*>-i W 41 U
3 U c M >.
J3 C. — C J3
« 41 41 -5 *
* — . JS — 41
.0 *J «, C
j= « a
0 W. -0 J3 41
3 M C J
*l J3 3 <«
00*0
4P 41 * 0 ll
^- J= 0
O W e 41 JS
* O ^ W
*4 W r- —
• W JC r4 M M
M J2 00  ja c
'V U W fc* •*• O -»*
*J ^M O. L< W U
1 -S S ^ S =-5
* 41 M C
O E 4* — * U — U
.CM 0 •-. C C w O
A »- - 3
<*- *J «M 41
4) m V 4» -• U
t» 41 *J rt 0 £
-0 C JS j= 0 U
•"*  ^ -• •*,
o w « o M
JZ — = J3 = I- O
tno £ M 3; o z










A

«n








—



 "O
•a- §
C M 0
«Q 41 U
b «
e <«
O 41 O
— CO
W CQ
u <— —
4J O
t, «j o
C 0 U
i !U
U, V- 41
M O
M *- -*
U 41 U
41 *J> 4J
 "-
c e ee
^ c .c
M e
3 b C.
X CiO





A




A

«n
*"
i





41 * O
— J= 0
•^ — ~
— «* o
a: c j=




W W «J
M 41 O JC
i as buildings, mil
>rotrudes above th
the inlet probe,
ilr 2 of AppiMtilf x
O -o a; —
= e = t~
o ^ ^*
b — ^
<» * — —
U 0 ^ ~
•• C a;
w r«. 5 ^
- IA CM J
J3 »—
o ° rf
1 C^r
e. 3~ g
"= - * 5
S «-i-ls
w Jo - C u
o £ > -*•
U Q. s — SC
Q M — f CB C (5
JS — C S ^3 en =.





A




A

«n
—
1






zn




n
O

„-
X
MM "c C
04 ft. W V
C T3 C. CL
--J -3 0.
T3 ^- *J • •<
— *J O
— 0 — «—
3 U - o
ja c. —
' M 4f 4t •
« •— — CO 41
0 _» C —
"6 w -^ ^ "S
S^ §=£
o or
4) 4) •) -•
owe •>.
« o •— *
w u (—01-
J= 00 U
3 — 3 -S J=
o- a — w
O jC — a -«
Si — o
-= "3 4 3 41
O J3 — i
— o «; c- o '«
^ i O 0 »
E— • » w « ^5 O
C O 4; C 1-
o e o — u w
A £ ~" -Z = ~" §

O C W S 0 ftC
~ = — J= O C
C is v» o s: "~ i
« i= 1'^ = o tS"





A




A

»n
—
i






^




o01
•j-


-------
                                               Section No.  2.0.2
                                               Revision  No. 1
                                               Date July 1, 1979
                                               Page 8  of 9

       3.    Plug the ports on  the manifold when  sampling lines are
  detached.
       4.    Maintain . a  flow  rate  in  the manifold that  is  either 2
  to  5 times the  total  sampling requirements or  at a  rate  equal zo
  the  total  sampling  requirement  plus  140 £/min  (5 ft: /mir-}.
  Either rate will help to  reduce the  sample residence  time in the
  manifold and  ensure  adequate gas  flow to the  monitoring instru-
  ments.
       5.    Maintain the vacuum in the  manifold £0.64  cm (O-25 in.)
  water gauge.   Keeping the  vacuum  low will  help to prevent the
  development of leaks.
  2.4  Support Services
       Most  of  the  support  services necessary for the  successful
  operation. of  ambient  air monitoring  networks can be  provided by
  the laboratory.   The major support  services are the  generation of
  reagent water  and the  preparation of standard atmospheres  for
  calibration of  equipment.    Table  2.3 summarizes guidelines  for
  quality control of these two support  services.
          In  addition to  the information presented above,  the
following should be considered when designing a sampling manifold:
     •    suspending  strips  of paper  in front  of the  blower's
          exhaust to permit a visual check of blower operation,
     •    positioning air conditioner vents away from the manifold
          to reduce condensation of  water vapor in the manifold,
     •    positioning  sample  ports  of  the  manifold toward  the
          ceiling  to • reduce  the  potential  for accumulation  of
          moisture in analyzer sampling lines,  and
     •    using  borosilicate  glass,  stainless  steel,  or  their
          equivalent for VOC sampling manifolds at PAMS sites is to
          avoid adsorption and desorption reactions of VOC' s on FEP
          Teflon.
Also, probe-siting criteria for Pb and PM10 NAMS and SLAMS for PAMS
have been added to 40  CFR 58 since the publication of this section
in 1979.

-------
                                             Section No. 2.0.2
                                             Revision No. 1
                                             Date July 1, 1979
                                             Page 9 of 9
Table  2.3  TECHNIQUES  FOR QUALITY CONTROL OF  SUPPORT SERVICES*
  Support
  service
            Parameters
         affecting  quality
     Control
    techniques
Laboratory
 and cali-
 bration
 gases
     Purity  specifications  -
     vary  among manufacturers

     Variation between  lots
               Atmospheric interferences
               Composition
 Develop purchasing
 guides

 Overlap use of old
 and new cylinders

 Adopt filtering .and
 drying procedures

 Ensure traceability-to
 primary standard
Reagents
 and
 water
     Commercial source
     variation
               Purity requirements
               Atmospheric interferences
               Generation and storage
               equipment
 Develop purchasing
 guides.   Batch test
 for conductivity

 Redistillation,  heat-
 ing,  deionization
 with ion exchange
' columns.

 Filtration of ex-
 change air

 Maintenance schedules'
 from manufacturer's
 recommendations  and
 from method require-
 ments
2.5  References
     4.

     5.
Guidelines of  Technical  Services  of a State Air Pollu-
tion  Control Agency.   Environmental Protection Agency.
Contract No. 68-02-0211.  1972.

Unpublished  research,. California  Air  Resource Board.
January 1977.

Field  Operations Guide  for  Automatic Air  Monitoring
Equipment.  Environmental Protection Agency.   APTD-0736
October 1972.

40 CFR 58, Appendices C,  D,  and E,  May 10, 1979.

Quality Control  Practice in  Processing  Air  Pollution
Samples.  Environmental  Protection Agency.   APTD-1132
March 1973.

-------
V

-------
                      Addendum to Section 2.0.3
                      Data Handling and Reporting
In addition to data validation information presented in this section, an excellent
document concerning this topic, Validation of Air Monitoring Data (EPA 600/4-
80-030), has been published by EPA.  It is available as document number PB 81
112534 from the National  Technical Information Service, 5285 Port Royal
Road, Springfield, Virginia 22161.

-------

-------
                                              Section No. 2.0.3
                                              Revision No. 1
                                              Date July 1, 1979
                                              Page 1 of 13
 3.0  DATA HANDLING AND REPORTING
 3 .-1.  Data Recording
      This  section  presents   standard  forms  generally  used  to
 record data .gathered  by ambient  air  monitoring systems,  and  it
 identifies problems common to data reporting.
 3.*1-1  Standard Forms for Reporting1 - All data  forms  should in-
 clude station identification  adequate  for tracing to the original
 site  description..  The site description should include  all  infor-
•mation identified on the SAROAD, form,  Figure  3.1;  the  same codes
 should be  assigned to  the SAROAD daily  and  hourly data  forms,
 Figures  3.2  and 3.3,  respectively.   These standard  EPA  data forms
 are designed to aid in transmitting data  from  the State and local
 agencies  to  the EPA  data bank in accordance  with 40 CFR  51,
 Part  51.7.   The detailed procedures to be followed in  completing
 these forms are . given on Reference 2.   These 'forms will  be up-
 dated • periodically  and • the user  should  maintain current  forms.
 If  computer  techniques  'are  used  for  recording   results,  the
 computer  system must  be designed  to  maintain compatibility be-
 tween the  SAROAD station  codes and  the  codes used by the computer
 program.   Whenever  station parameters  change or when  a station
 is  moved,  an updated site identification  form  (.Figure 3.1)  should
 be  submitted to the data bank.
      Identification  errors  can  be avoided  by  preprinting all
 forms,  checklists,   calibration   forms,  and so   forth,   with the
 station  identification.   'If  this  technique is  adopted, control
must  be  employed to be  certain that unused  forms  are   discarded
 and-new  ones  printed  when the  station  identification  changes.
Preprinting  the pollutant I.D.  and the proper decimal points for
that  pollutant on  the reporting  forms  can eliminate the problem
of misplaced decimals.
     Acceptability  limits  for start-stop times, flow  rate,  and
other  routine   system  checks   performed by  the  operator should

-------
                                                        Section  No.  2.0.3
                                                        Revision No. 1
                                                        Date  July 1, 1979
                                                        Page  2 of 13
                         ENVIRONMENTAL PROTECTION AGENCY
                           National Aerometric Data Bank
                          Research Triangle Park. N. C. 27711

                          SAROAO Site Identification Form

                                          6/6/77
                       D

TO 8E COMPLETE1" BY THE
i A\ C^ Q. \ i

• UO4< 3 characters)
Bj ^
ernc»v-di»oo
•"•»'• County Name (15
City Population (right justified)
|OlO|0|3!2|5l5ll
JJ 43
1 0 0 1 W
(U fl tl
UTM Zone
•1 1 1


(O| 'M^l' IWI
characters)
54 IS i« 4/ 58 49
Longitude Latitude
Deg. Mm. Sec. Oeg. Mm. Sec.
I ll7(4l3lZk
ft fi S4 «b 6; 63
Easting Coord., meters
1
D N|3l4|/lolo|Sl
69 ."'• /I ;: ;3 M »5 18
Northing Coord., meters
1 1 1 1 1 1 1

• ^\ • f"3
rKiiou riiv* HcSot
Supporting Agency
j^^,- Rr-rtir^
(61 characters)
Supporting Agency, continued


"':' Optional, Comments that will help identify
the sampling site (132 characters)

irn







IF)


DO NOT WRITE
State Area
A
1 7 a < 4 6
Agency Project
a
HERE
Site
1 1
r a 9 10
II 12 13
Region Action
a a
') 80
State Area
8 i
Site
|
1 .' 3 » S G 7 8 9 JO
Agency Proiect SMSA Actio
L 1 1 U !_'
i — i i — i i i i
11 1*13 1415
State Area
c 1
1 7 3 4 S 6
Agency Project
n ED
11 1? 13
State Area
0 |
1 .-3*56
Agency Project
n ED
! i 17 13
State Area
E
n n
i ..i 	 1 1 — i
16 i; in
Site
F
7 a s ..
Action
n
ao
Site
1
/ 8 9 10
Action
n
30
Site
1
           Abbreviated Site Address (25 characters)
OMB No. 1S8-R0012
Approval expires 6/30/76
        3  «   5  6  '   89
Agency  Project      Action

 D  ED       D
                                   (over)
       Figure 3.1   SAROAD site  identification  (front).

-------
                                                              Section No.  2.0.3
                                                              Revision No.   1
                                                              Date  July  1,   1979
                                                              Page  3  of  13
            SAROAD  Site  Identification  Form  (continued)
 TO BE COMPLETED BY THE REPORTING AGENCY
                                                    DO NOT WRITE HERE
 (R IO5O  San
            Sampling Site Address (41 characters)
 Check the ONE
 major category that
 best describes the
 location of the
 sampling site.
 1.CH CENTER CITY
   Address, continued

     Next, check the subcategory
     that best describes the domi-
     nating influence on the sampler
     within approximately a 1-mile
     radius of the sampling site.
2.12SI SUBURBAN
                    X
     RURAL
        1. Industrial
        2. Residential
        3. Commercial
        4. Mobile
        1. Industrial
        2. Residential
        3. Commercial
        4. Mobile
        1. Near urban
        2. Agricultural
        3. Commercial
        4. Industrial
        5. None of the above
4.f~l REMOTE
Specify
units	
20-Pi-
             Elevation of sampler above ground
Specify
units 	
         Elevation of sampler above mean sea level

Circle pertinent time zone:   EASTERN   CENTRAL

MOUNTAIN   PACIFIC   YUKON   ALASKA  BERING
HAWAII
                                                     State
                                                 Area
                                                             Site
                                                    1   234561
                                                    Agency
                                                     n
  Project
 an
Station Type
                                                                 County Code
                                                               AOCR Number
                                                                61  63  63
                                                                   AQCR Population
                                                                      69  70   71
                                                            Elevation/Gr
                                             J3  73  74
                                                   Elevation/MSL
                                                     Time
                                                     Zone   Action
                                                   75  76  77   78
                          Figure 3.1  (continued)
              SAROAD site  identification  form  (back).

-------
                                                       Section No.  2.0.3
                                                       Revision No.  1
                                                       Date  July  1,.1979
                                                       Page  4  of  13
                         ENVIRONMENTAL PROTECTION AGENCY
                            National Aerometric Dad Bank
                          Research Triangle Park. N. C. 27711

                             SAROAD Daily Data Form
24-hour or .greater sampling interval
OMB No. 158-R0012
Approval empires 6/30/76
2
Cali-rorKii'CL Air Re5oufa?^ Boovei
' l4i«UI
Units
lo
1

31' 31
34 3i 36
n
I



















— —I 	
'


1
1
'

i

O


















^J"


















-f -
j














DP
o]
32

Time Interval
Name
PARAMETER
Code

Me





3t 39 4-J Jl
hod . Units DP
nnnn
a] 44 4^ 4€


















































....





































































• Ag
[
Slate Area Site •
0|Sl©|4-|4.k
Dlolokvl
; j 4 •. 6 ; i u in
ency Project Tirre Year
II 1. 13 14 IS 16
Name
PARAMETER
Code
C
Me
J)
1
hod
D
iS
6'



















-"












Units
| i
S? 91 t-J
















































.





































DP
n
6U

Month
lolel
U II
Name
PARAMETER
Code

Met
c





66 6/68 (9
hod .Units DP
nmn
7t 77 ;j /4
'') '6 •' IB




















.....



































































































                               1)10
                                            4371
              Figure  3.2    SAROAD  daily  data  form.

-------
                                          Section No. 2.0.3
                                          Revision No.  1
                                          Date July 1,  1979
                                          Page 5 of 13
        SSS
       •steU
 J£
  fQ
  •Q-
  CD

  fe
i-s 5
    £.„. ty
        = d_l
         C.3
II
    a
    o
        e C
        H
 ~3
-eg
-  u
I,C
5 S

 O

"*c
i3
< a
? 'C
1/1 C
 i
         OCi-s
               nccc
= 3
                0^
           C-
           OCC
                CCQ
           CC
           GO
                CC
              00
                 u
                 CC
                        •r
                       —;~M
                      -HH
                       i
                         ,-.-L
                          4-r1
                              i
                                 H-
                                --h
                                  rn
                                 ^•fi-H
                                     r^--V-^4
                                   -1-U l.
                                        i n^r
                                        THTT
                                     u
                               11
                               '  i
                                  j	i
                                          -t-r
                                           l
                               trur
itn
-f
                                  ~tt!

-------
                                              Section No.  2.0.3
                                              Revision No.  1
                                              Date  July 1,  1979
                                              Page  6  of 13

 appear on the data recording  form  as  a reminder to  the operator.
 If a  value  outside  these  limits  of  acceptability  is recorded,
 the  operator should flag the  value  for the  attention of  individ-
 uals  performing  data  validation functions.
 3.1.2  Data Errors in  Intermittent  Sampling -  The  most common
 errors in recording data  in the field  are transposition of digits
 and  incorrect placements  of  decimal  points.  These errors are
 almost impossible  to  detect.   The decimal error can  be avoided to
 some  extent by  providing an  operator  with  the  following guide-
 lines for accuracy:
                                    • Number  of     Example cone:,
 Pollutant                      '   decimal places      pg/m3	
 Suspended particulate matter               0          87
 Benzene soluble  organic"matter            1'           6.1
 Sulfates                                   1          10.1
 Nitrates                                   1           2.3
 Ammonium                     .         .     1      .     0.7
 Sulfur dioxide                             0          98
 Nitrogen  dioxide                           0          40
 Nitric oxide                               08
 Carbon monoxide                            1           4.5
 Total  oxidants                             0          100
 Total  hydrocarbons                         1           2.7
 Ozone          •                           0           72
 Methane                              '      1            ]_ _ 5
 3.1.3   Data  Errors, in Continuous Sampling  -  Data  error's  in  con-
 tinuous  sampling  primarily include errors  in recording device
 functioning, and errors in strip chart reading for manual techni-
 ques  or  in  data transmission  for.  automated techniques  of  data
 recording.   See  Sections  1.4.10 and 1.4.17  of Volume I  of  this
Handbook  for additional information concerning data reduction and
data  validation.   In particular,  for  errors in  automated tech-
niques  of data  transmission   and recording,  see  Section 1.4;10.

-------
                                              Section No.  2.0.3
                                              Revision No.  1
                                              Date July  1,  1979
                                              Page 7 of  13

      Strip chart errors - Errors in mechanical functions  of strip
charts  can  occur.   General guidelines to avoid errors  or  loss of
data  caused by mechanical problems follow:
      1.   Perform  a daily check to  assure  an adequate supply of
strip chart paper.  Check  the ink-level in the  recorder pen to
verify  that the .level  is  adequate for  the  next sampling period
and that the pen tip is not blocked.
      2. 'Perform  a daily  check to  verify  that the pen on the
recorder  aligns  with the baseline of the strip  chart during the
instrument zero check.
      3.   Verify the  timing of  the  strip. chart drive  against a
standard timepiece  immediately after  installation of the recorder
and at  intervals dictated by experience with  the recorder.
      4.   Replace  recorder pens,  and soak  in  cleaning solution
occasionally.
      5.   Examine  the  strip  chart  for apparent evidence of chart
drag  of malfunction, and mark suspected intervals.
      When reviewing a strip chart, typical  signs  of system mal-
function are:
      1.   A straight trace  for several hours (other than minimum
detectable).
      2.   Excessive noise as indicated by a wide solid trace, or
erratic behavior such as spikes.that are sharper than is possible
with .the normal instrument  response  time.   Noisy- outputs usually
result when analyzers are exposed to vibrations.
    .3.   A long steady increase or decrease in deflection.'
     4.   A  cyclic pattern  of  the  trace with  a definite  time
period  indicating   a "sensitivity  to   changes  in temperature  or
parameters other than the pollutant concentration.
     5.   Periods  where the  trace  drops  below the  zero baseline.
This  may  result from  a larger-than-normal  drop  in  the  ambient
room temperature or power line  voltage.
     Void any data  for any time interval for which  malfunction of
the sampling system is detected.

-------
                                              Section No.  2.0.3
                                              Revision No.  1
                                              Date July 1,  1979
                                              Page 8  of 13

      Suggestions for  minimizing errors  in reading  strip charts
 are as follows:                                     '
      1.    Chart  readers should  be  trained with  a standard  strip
 of chart,  whose readings  have  been  determined by  one  or  more
 experienced readers.   When the  new reader  can perform adequately
 on the  standard strip, then permit him/her  to read new sample
 charts.   A suggested  procedure for  reading  a  strip chart  is  given
 in the following subsection.
      2.    An  individual  should   spend  only  a portion of  a day
 reading  strip  charts  since  productivity  and  reliability are
 expected to decrease  after  a  few hours.
      3.    A senior  technician should check at  least  5% to 10% of
 the strip' chart  values  reduced.  The percentage of checking is
 arbitrary,  and depends on both the depth  of training  of the  junior
 technicians  and  the  time  available  for  checking their work.  If
 minimum  performance criteria  established  for  a  particular network
 are not  being  met,  additional  training is indicated.
      4.    Use  a  chart re-ader to  reduce technician fatigue and to.
 improve  accuracy in data reduction.
      Data  reduction from strip chart  - To  obtain hourly average
 concentrations  from  a  strip  chart record, the following" proce-
 dures  are  used:
      1.    Be sure the strip chart record for the sampling period
 has span and zero traces at the beginning  and at the end of the
 sampling period.
      2.    Fill in the identification data  called for at the top
 of  an hourly averages form, Figure 3.4.
      3.   Using  a straight  edge,  draw a line from the zero base-
 line  at  the  start of the sampling period to the zero 'baseline at
 the end  of the  sampling period.  This line  represents the  zero
baseline to be used for the sampling period.
     4.   Read the  zero baseline in percent of chart at the mid-
point of each  hour  interval,  and record the value on Figure 3.4.
     5.   Determine the  hourly  averages  by  using a transparent
straight edge  at least 25  mm (1 in.) long.   Place  the straight

-------
                                                   Section No. 2.0.3
                                                   Revision No. 1
                                                   Date July 1, 1979
                                                   Page 9 of 13
CITY.
SITE
      ^
                 Cc
LOCATI ON IOSO Sar\
                             RA
SITE N0._
POLLUTANT.
OPERATOR
                                              V
DATE
*j£






	












	





	
HOUR
00











""








	







READING ! ZERO BASELINE 1 DIFFERENCF 1 ADD + 5 1 PPm
ORIGINAL
:.JL






























CHECK































ORIGINAL
S~






























CHECK































ORIGINAL
¥






.























CHECK































ORIGINAL
p






























CHECK































ORIGINAL
0. 0&0






























CHECK d

	 _

























	 —

The check  columns are for use by  an independent audit.
performed  on  approximately 7 percent -of the data.
                                                           The check is
        Figure 3.4  Sample form for recording hourly averages.

-------
                                              Section No.  2.0.3
                                              Revision No.  1
                                              Date July 1,  1979
                                              Page 10 of 13

 edge parallel to  the horizontal chart  division lines.   For  the
 interval  of interest between two vertical  hour lines,  adjust  the
 straight  edge between the lowest and  highest points of the trace
 in thac interval,  keeping the straight edge parallel to the chart
 division  lines until   the total area   above  the  straight edge
 bounded by the trace and the hour lines  is  estimated to equal  the
 total  area below the straight edge bounded, by the  trace  and hour
 lines.  Read  and  record-the percentage of chart deflection on  the
 hourly  average  form.   Repeat  the  procedure  for   all  the hour
 intervals  sampled which have not  been marked invalid.   Record
 all  values in the  column headed  "Reading -  Original" (Figure 3.4).
     6.    Subtract  the   zero  baseline  value  from  the   reading
 value,  and record  the difference (Figure 3.4).
     7.    Add .the percentage of zero  offset, +5, to the  differ-
 ence.
     8.    Convert  the percentage  chart values  to  concentration
 (ppm) ' using the most recent  calibration curve.   Record the  ppm
 NO2  values  in the  last column (Figure  3.4).
     Data  validation  -  The  purpose   of data  validation  is-  to
 detect  and then  verify  any data  values that  may not represent
 actual  air  quality conditions at the sampling station.  Effective
 data validation  procedures usually are  handled completely inde-
 pendently  from the procedures  of initial data collection.  More-
 over, it  is  advisable that the  individuals responsible for data
 validation not be directly involved with data collection.
     Both manual and computer-oriented systems require  individual
 reviews of  all data  tabulations.   As  an individual  scans tabula-
 tions,   there  is  no  way  to determine  that  all  values are valid.
The  purpose of manual inspection  is  to spot unusually high  (or
 low) values that, might  indicate a  gross error  in the data col-
lecton  system.   Obviously, to recognize that  the  reported con-
centration  of a  given pollutant is extreme,  the individual must
have basic  knowledge of the major pollutants  and of air quality
conditions  prevalent   at  the- reporting station.    Data values

-------
                                              Section No. 2.0.3
                                              Revision No.  1
                                              Date July 1,  1979
                                              Page 11 of 13

 considered questionable are referred to the data collection group
 for verification.   This scanning for high/low values is sensitive
 tc spurious extreme  values but not  to intermediate values  that
 could be grossly in error.
      Manual review of  data tabulations also  allows  detection of
 uncorrected drift  in  the  zero baseline  of a continuous -sensor.
 Zero drift may be  indicated when  the daily minimum concentration
 tends to  increase  or  decrease  from  the  norm  over a period of
 several  days.   For  example,  at  most sampling stations  the  early
 morning   (3:00  a.m.  to  4:00 a.m.)   concentrations   of   carbon
 monoxide tend  to reach a minimum  of  2 to 4 ppm.   If the  minimum
 concentration  differs significantly  from  this,  a  zero  drift  can
 be suspected.    Zero  drift could  be confirmed  by  review of  the
 original strip chart.
      In  an automated data processing system,  procedures for data
 validation can  easily  be  incorporated into  the basic  software.
 The  computer  can . be programmed  to. scan each data value before
 preparing an  output tabulation.   Questionable data  values  are
 then  flagged on the data tabulation to  indicate  a possible error.
      A computer  can easily handle inspection  for extreme values.
 Tables of high and  low values for each pollutant at each sampling
 station  can be  built into the  program.   The  check for extreme
 values can be  further refined to account for time of day, time of
 week, and  other cyclic conditions.
      Because   the  computer  can  perform  computations   and  make
 comparisons extremely  rapidly,  it can  also make some determina-
 tion  concerning  the validity of data values  that are not neces-
 sarily high or low.   Data  validation procedures should be recom-
mended as  standard  operating procedures.-"2   One way to do this is
to test  the difference  between  successive data values,  since one
would not normally expect very rapid changes in concentrations of
a  pollutant during  a  5-min or  1-h  reporting period.   When the
difference  between  two  successive  values  exceeds a  predeter-
mined value, the  tabulation can be flagged,  with an- appropriate

-------
                                            • Section No. 2.0.3
                                             Revision No. 1
                                             Date July 1, 1979
                                             Page 12 of 13

symbol.   For example,  if two  values for  hourly  sulfur dioxide
                                                  3
concentrations   differ  by   >0.05 ppm  (100 pg/m  ),   the   data
wlidd'cion  stall may wish  to recheck the  original records from
which the data were obtained.
i. 1  Systematic  Data Management
     An  orderly  process of  data  management based  on analysis of
all the  data-handling procedures  and their interrelationships is
sometimes  called a "systems"  approach.  This  kind of systematic
overview  of  the total  data function  is   accomplished  in  three
phases:
     1.   Surveying  current  and   future reporting requirements.
     2.   Outlining the present  routine flow  of data within and
outside the agency.
     3.   Redesigning  the  current system  to  allow maximum func-
tional overlap of filing and  retrieval routines.
     A survey of current  reporting  requirements involves summa-
rizing and  categorizing the reports .currently required and their
important  data  elements.   The purpose  of  this analysis  is  to
identify  report  elements  that require  similar input,  to  allow
optimum scheduling, and to differentiate between required 'reports
and those  provided as  a  service.  Future  reporting requirements
will be based on projected legal  requirements,  projected develop-
ments of systems for communicating  with various data banks,  and
projected growth of the air quality surveillance network.
     Outlining present  data  flow  requires  a review of the origin
of each  data form, the- editing procedures applied, the calcula-
tions performed,  the  application  of quality control procedures,.
and the  reports  for  which  each   form  is  used.  The. purpose  of
outlining  the data flow  is  to "identify data  elements  that  are
subjected to similar checks and to similar calculating procedures
and to classify  them "according to their points of origin.   Once
again,  this procedure  provides  a  means of preventing unnecessary
duplication.

-------
                                              Section No.  2.0.3
                                              Revision No.  1
                                              Date  July 1,  1979
                                              Page  13 of 13

     As  a  final step  in  systematic  data management,  the  data
system  should be  continually updated.  The  following items  are
suggested  for review:
     1.    What operations  are duplicated in the  system?
     2.    Blow  can the  system be  changed to eliminate  needless
duplications?  .
     3.    How  do  the . manual  systems  and  computerized   systems
augment each other?
     4.    Are  the  data formats,  identification codes, and  other
elements compatible throughout the system?
     5.    Can  reporting  schedules  be changed  to  minimize  the
filing and retrieval of each data record?
     6.    Can  special  techniques, such as the  use of multipart
forms/  be  applied to minimize data transposition?
     7.    Are filing and  retrieval  systems sufficiently  flexible
to allow expansion or upgrading at minimum cost?
3.3  Evaluation of Air Quality Data   •   .            •
     Minimum  requirements  and procedures  for .evaluation of"  air
quality data  are given in References  2, 3,  and 4.  However,  the
analyst  should  also   use  Reference 4  for specific  statistical
methodology  for  analysis  of  air  quality  trends.   A bacic did
CUEn-inn nf 1-hn r-t-i-M r-h-i r--n rr^-t-fr^r; -j n contained in th~  appendices
to Volume  I of thic Handbook.
3.4  References
     1.   Quality Control  Practice  in Processing  Air Pollution
          Samples.   Environmental  Protection Agency.  APTD-1132.
          March 1973.
     2.   AEROS  Manual  Series Volume  II:   AEROS  USER'S  Manual
          EPA-450/2-76-029, OAQPS  No.  1.2-.039.   December 1976.
     3.   Guidelines  for  the  Evaluation  of  Air  Quality  Trends.
          Environmental  Protection  Agency,   Research  Triangle
          Park,  N.C.   OAQPS No.  1.2-014.  December 1974.
     4.   Guidelines   for   the  Evaluation  of  Air Quality Data.
          Environmental  Protection  Agency,   Research  Triangle
          Park,  N.C.   OAQPS No.  1.2-015.  January 1975.

-------

-------
                                              Section No.  2.0.4
                                              Revision No.  1
                                              Date July 1,  1979
                                              Page 1  of 6
 4.0  REFERENCE AND EQUIVALENT METHODS
      For  monitoring   in  a  SLAMS   or  NAMS  network,   either
 reference  or  equivalent methods  are  generally required.3   This
 requirement  is specified  in-40  CFR Part 58, Appendix C  (Federal
 Register,  Vol  44,  May  10,  1979,  page  27584). •  In  addition,
 reference  or equivalent  methods  may  be  required  for  certain
 other   monitoring    applications—for    example,    monitoring
 associated  with  prevention  of  significant  deterioration  (PSD).
 Requiring  the uses  of .reference or equivalent  methods  =helps to
 assure   that   air quality  measurements ' are  made with  methods
 which have  been shown to have adequate  accuracy and reliability.
      The   definitions   and   specifications  of   reference  and
 equivalent  methods  are  given in  40  CFR Part  53.  However, for
 most jnonitoring  applications  the distinction  between  reference
 and   equivalent   methods   is  unimportant;,  either, may  be   used
 interchangeably..
      Reference  and  equivalent  methods  may  be either manual or
 automated  (analyzers).   For  S02,  particulates,  and  Pb,  the
 reference  method   for each  is  a  unique  manual  method  that is
 completely  specified  in  an Appendix  to .40  CFR Part  50;  all
 other  methods  for  S02  and  Pb  qualify  as equivalent  methods.
 (As   yet,   there   is  no   provision  in   the   regulations  for
 designating equivalent  methods  for particulates.) • For  CO, NO2/
 and   O3/.   Part   50   gives   only  a  measurement  principle'  and
 calibration procedure applicable to reference  methods for those
pollutants.   Automated  methods  (analyzers)  for  these  pollutants
may  be  designated  either  as  reference  methods  or  equivalent
methods, depending  on  whether  or not  the  methods utilize the
same  measurement  principle  and  calibration procedure specified
in Part 50  for  reference  methods. ' since  any  analyzer which
a
  Certain  exceptions, to'  this  general  requirement  are  provided
  in Appendix C of 40 CFR Part 58.

-------
                                             Section No: 2.0.4
                                             Revision No. 1
                                             Date July 1, 1979
                                             Page 2 of 6

meets  the  requirements of  the specified  measurement principle
and  calibration  procedure  may  be  designated  as  a  reference
method,  there are  numerous reference  methods for CO,  NO-, and
Og.  Further information on  this  subject  is  in  the preamble to
40  CFR Part 53,  published  in  the  Federal Register,  February 18,
1975 (40 FR  7045-7046),..
     Except   for  the   (unique)   reference   methods   for  S02/
particulates,  and  Pb  specified  in Part  50,  all  reference and
equivalent  methods  must be officially  designated as  such by EPA
under  the   provisions  of  40  CFR  Part   53.   Notice  of  each
designated  method is  published in  the Federal  Register  at the
time   of  designation.   In  addition,  a   current  list  of  all
designated  reference  and  equivalent  methods  is  maintained and
updated by  EPA whenever a  new method  is  designated.   This list
may  be obtained  from  the  Quality Assurance  Coordinator  at any
EPA  regional   office  or  from   the  Environmental  Monitoring
Systems  Laboratory,   Department   E,   MD-77,  -Research  Triangle
Park,  North  Carolina  27711.   Moreover,  any analyzer  offered
for  sale  as  a reference  or  equivalent  method  after April 16,
1976,  must  bear  a label or sticker  indicating that the analyzer
has  been  designated as a reference  or  equivalent method by EPA.
     For  automated  methods,  a  designation applies  only  to  an
analyzer  which  is   identical  to  the  analyzer described  in the
designation.   Since in the past manufacturers may  have changed
or  modified analyzers  without  changing the  model number,  the
model  number  alone  does  not  necessarily  indicate  that  an
analyzer  is   covered   under   a - designation.    In  many  cases,
analyzers manufactured  prior  to the  designation  may  be' upgraded
(e.g.,   by  minor  modification  or  by  substitution  of  a  new
operation or instruction  manual) to  make  them identical  to the
designated method  and thus to  achieve the designated  status  at
modest cost.   The manufacturer should  be  consulted  to determine
whether an  analyzer  is covered under  a designation  or whether,
it is feasibile to upgrade it to the designated status.

-------
                                              Section No.  2.0.4
                                              Revision No.  1
                                              Date July 1,  1979
                                              Page 3 of 6

 Furthermore,   any  modification  to  a  reference  or  equivalent
 method made by a  user  must be approved by EPA  if  the designated
 status is  to  be  maintained  (see  Section 2.8 of 40  CFR Part 58,
 Appendix C).
      Sellers   of  designated  automated methods  must comply  with
 the conditions summarized below:
     . 1.   A copy  of the approved operation or instruction  manual
 must  accompany  the  analyzer  when   it   is delivered  to   the
 ultimate purchaser.
      2.   The  analyzer  must not generate any unreasonable  hazard
 to operators or to the  environment.
      3.   The  analyzer  must function . within the  limits  of  the
 performance specifications in Table 4.1  for  at  least 1 yr after
 delivery when  maintained  and  operated  in  accordance  with  the
 operation manual.
      4: '  Any   analyzer   offered  for   sale   as   a   reference  or
 equivalent  method must  bear  a label  or  sticker indicating 'that
 it has  been  designated  as  a reference or equivalent method in
 accordance  with Part 53.
      5.   If such an analyzer has one  or more selectable ranges,
 the  label  or  sticker  must be  placed  in close  proximity  to  the
 range  selector and must  indicate  which  range  or  ranges, have
 been designated as reference or equivalent methods.
     6.    An   applicant   who   offers   analyzers   for  sale  as
 reference  or  equivalent methods is  required  to  maintain  a list
 of ultimate purchasers  of  such analyzers  and  to  notify them
 within 30 days if  a  reference or equivalent method designation
 applicable  to  the analyzers has been  cancelled  or if adjustment
 of the analyzers  is necessary  under 40 CFR  53.11(b) to  avoid a
 cancellation.
     7.   An   applicant   who   modifies  an  analyzer  previously
 designated  as   a  reference  or  equivalent  method  is  neither
permitted  to  sell  the  modified   analyzer  as   a  reference  or
 equivalent method  (although an applicant may choose to sell  it

-------
 0)
rH
XI
d) ro
&VH
0 X
&H O
•P BrH
•H 13
55
(U
O -H
A X
H O
(0 G
0 0
£

G
O
N
O



0


0
in
•
o


in
o
o
•
0



in
0
0
*
o




£
ft
ft





















•H
jj
Q1
(U

0)
U
G
0)
i-(
cu
M-J
in


-p
•H
£
-H
rH

0)
Cn

m
(_i

j^
0)
ft
ft
3

*H
O

o^5
0
CM



O
• o m in
in CM rH rH
+ 1



in
• o in in
CM rH
+ 1



o
• o in m
in CM rH rH
+ 1




o
• o m in
in CM rH rH
+ 1
'

G G G
•H -H -H
*z G e G

p
•H
£
•H
rH

0)
Co
G

!H

[.4
(U
ft
ft
 W rH H

CM
O
•
0



in
•
0



rH
O
•
O




rH
O
.
o




£
ft
ft

-P
-H
£
•H
rH

• 
-------
                                              Section No.  2.0.4
                                              Revision No.  1
                                              Date July 1,  1979
                                              Page 5 of 6

 without such  representataion)  nor  permitted to  attach  a  label
 or sticker  to the  analyzer (as  modified)  under  the  provisions
 described above,  until the applicant  has  received notice  under
 40 CFR  53.14(c)  that  either  the original  designation or  a  new
 designation  applies  to  the  method  as modified or  until  the
 applicant  has  applied   for   and  received  notice   of  - a  new
 reference or  equivalent  method  determination  for the  analyzer
 as modified.                     .
      Aside   from   occasional   breakdowns   or   malfunctions,
 consistent  or   repeated   noncompliance   with   any   of   these
 conditions  should  be   reported  to  EPA  at  the  address   given
 previously.
      In   selecting  designated   methods,    keep   in  mind  that
 designation  of  a method  indicates only  that  it meets  certain
 minimum  -standards.   Competitive  differences  still  exist  among
 designated  analyzers.    Some   analyzers  or  methods  may  have
 performance,   operations,  economic,'  or .other  advantages  over
 other  analyzers  or  methods.   Thus   the   need   for   a  careful
 selection   process   based   on  the   individual   air   monitoring
 application   and   circumstances    is    still    very    important.
 However,  some  of the performance  tests and  other criteria used
 to  qualify a  method  for  designation   as a  reference  or equiv-
 alent  method  are intended only  as pass/fail  tests to determine
 compliance  with  the  minimum  standards.   Therefore,   test data
 from  such  tests may not. be  usable  to  quantitatively compare one
 method  with  another  method.    Furthermore,  designation   as  a
 reference  or  equivalent  method  provides  no  guarantee  that  a
 particular  analyzer  will  operate properly,  since' any ,analyzer
 can malfunction.   So an  on-going  quality assurance  progaram  is
 necessary  and  required for  designated  methods  under  Appendix A
 for  SLAMS   and  Appendix   B   for   PSD  of   40   CFR  Part  58.
Appendices-  A  and  B  require   the  monitoring  organization  to
 establish  an   internal   quality   control   program.    Specific
guidance for  a minimum quality control program is  described  in
Section 2.0.9 of this  Handbook..

-------
                                             Section No. 2.0.4
                                             Revision No. 1
                                             Date July 1, 1979
                                             Page 6 of 6

     Many  organizations  have   elected   to  specify  'or  select
designated reference  or  equivalent  methods even  for monitoring
applications  other  than  those  required  by  Part  58.    This
practice may offer significant  advantages, such  as:   (1)  ease
of   specification,   (2)   guarantee   of  minimum   performance,.
(3) better instruction  manuals,  (4) flexibility of application,
and (5) increased credibility of measurements.
4.1  References
     1.   Code  of Federal  Regulations  40.   Protection of  the
Environment.   Parts 50  to 69.   As ammended, May 10,  1979.

-------
                                             Section No. 2.0.5
                                             Revision No. 0
                                             Date May 1, 1978
                                             Page 1 of 5
5.0  RECOMMENDED QUALITY ASSURANCE PROGRAM FOR AMBIENT AIR
     MEASUREMENTS
     In order  for  air monitoring data to be useful, they must be
of acceptable quality.  The dissemination and use of data of poor
or unknown quality can lead to incorrect decisions with regard to
environmental standards and regulatory actions.  The gathering of
air  monitoring data  under the  umbrella  of a  quality assurance
program does' much  to  avoid regulatory mistakes; thus all control
agencies  should  vigorously pursue  the  implementation of  such
measures.                              '
     The  major  elements  of  a  satisfactory  quality  assurance
program  are  (-1)  the   availability  of  an evaluated  measurement
methodology  which  is   adequate   for '  its   intended  purpose,
(2) satisfactory performance by  organizations  collecting the air
pollution monitoring data, (3) documentation of. quality assurance
practices,  and (4) the  .availability o.f  competent  technical as-
sistance  for organizations  needing  to improve  their performance.
It is imperative that the management of the monitoring program be
committed to  a quality  assurance program and that  adequate re-
sources be  available  to carry on the  activities  involved  in its
major elements.           •                      '
     To be s.ure, the implementation of a formal quality assurance
program does have  its  price,  but the cost of collecting good air
monitoring  data is far  less  than  the cost of making incorrect
regulatory decisions because of  poor data.   Experience has shown
that an agency should be  prepared  to spend between  10% and 25%
of its monitoring budget for its  quality assurance program.
5.1   Development and Implementation of Air Pollution Measurement
      Quality Assurance Programs
     Federal,  State, and  local agencies  all  have  important roles
to play  in developing and  implementing  a satisfactory  quality
a
 Recommendations  of  EPA's  Standing  Air monitoring  Work  Group
.(SAMWG)                                                         v

-------
                                              Section No. 2.0.5
                                              Revision No. 0
                                              Date May 1, 1978
                                              Page 2 of 5

 assurance program.  EPA's responsibility  is  to develop the tools
 needed to carry on  a  quality assurance program, and  it is  up to
 the State and"local agencies to implement their programs.
 5.1.1  Role of EPA Headquarters  -  EPA headquarters has the fol-
 lowing responsibilities:
      1.    To be certain  that the methods and  procedures used in
 making air  pollution  measurements  are well  evaluated  and  that
 their limits of precision and accuracy are well understood.
      2.    To determine the performance of laboratories making air
 pollution measurements of importance to  the  regulatory process.
      3.    To implement  satisfactory  quality  assurance  programs
 over EPA's air pollution monitoring  which have the potential for
 generating data used for  setting, standards.
      4.    To be certain that air monitoring data of importance to
 the regulatory  process are of satisfactory quality.
      5.    To share  with  EPA  regional offices - the rendering  of
 technical  assistance  to .the air pollution monitoring  community.
 5.1.2  Role of the EPA Regional Offices - The  major responsibil-
 ity of  EPA's  regional  offices  is  the coordination . of  quality
 assurance  matters  between  the  various  elements of  EPA and  the
 State and local agencies.   This role requires that the  regional
 offices make available to the State  and local  agencies  the  tech-
 nical  information  'and  quality assurance programs which  EPA  head-
 quarters  has developed  and   make  known to EPA headquarters  the
 unmet  quality assurance needs  of the State  and local  agencies.
 Another  very important  function of  the   regional  office is  the
 evaluation   of   the  capabilities  of  State   and  local  agency
 laboratories to measure air pollutants of  regulatory concern.  To
be effective in these'roles,   the regional  offices should maintain
 and  strengthen  their  technical  capabilities  with respect to  air
pollution monitoring.
5'1-3  Role  of  State and Local Agencies -  The major responsibil-
ity  of State and local agencies  is the  implementation of satis-
factory  quality assurance   programs  over  the monitoring  which

-------
                                              Section No.  2.0.5
                                              Revision No.  0
                                              Date May 1,  1978
                                              Page 3  of  5

yields  the air quality  data needed  for  the regulatory  process.
It  is the  responsibility of  State  and local  agencies  to implement
these programs in their  own laboratories and  in any  consulting
and contractor laboratories  which they may  use to obtain data of
importance to  the regulatory process.
5.2 Minimum Quality Assurance. Programs
     Comprehensive quality assurance  programs  in air  monitoring
are relatively new,  and many agencies responsible  for air moni-
toring  have  not  formalized their quality  assurance  activities
into  an identifiable  program.   As  an aid  to  agencies  who are
developing  quality  assurance programs and  to  agencies who wish
to  review  existing  programs,  we  present here  those  activities
considered to  be essential in an air  pollution monitoring quality
assurance  program.  These  essential  activities  and other aspects
of  a complete quality assurance program  are described in detail
in  "Quality  Assurance Handbook   for Air Pollution Measurement
Systems-- Volume I,   Principles"   (EPA-600/9-76-005),   thus  this
document  should be  consulted  in establishing  or  evaluating  a
quality assurance program.
     A  suggested  sequence for the development  of a quality as-
surance program is  given  below.   About  12  mo  is  required for
complete implementation.
5.2.1  Develop Immediately -
     1.    Agency quality  assurance policy   and objective - Each
agency  should  develop a  written.  quality assurance  policy,  and
this policy should be  made known  to  all  agency  personnel.  As  a
minimum, this  policy should create an awareness  of quality assur-
ance activities, provide  specific procedures for implementing  a
quality  control  program,  provide for   corrective   action  when
required,   state the  quality assurance objectives for each major
monitoring project operated  by the agency,  and  explicitly dele-
gate authority to implement  quality  assurance systems planned by
management officials.
     2.    Organization and  responsibilities -  An  organization
chart showing  the  key agency personnel and their areas of quality

-------
                                              Section No.  2.0.5
                                              Revision No.  0
                                              Date May 1,  1978
                                              Page 4 of 5

 assurance responsibilities should be prepared.   A quality assur-
 ance  coordinator  should  be  designated  for the  agency.   This
 designee should  be responsible  for  the coordination  of  quality
 assurance activities within  the agency and  with  other agencies.
 5.2.2  Develop Within Six  Months -
      1. '  Measurement method review and application  -  All exist-
 ing methods  (sampling and  analysis) used for  routine  measurements
 should be reviewed  and revised if necessary; written  procedures
 should be prepared where none exists.   A  document control system
 should be developed  for  these methods  to  keep  agency personnel
 abreast of  changes  in methodology.  Any  ambient air  monitoring
 for  criteria  pollutants  conducted  under State  Implementation
 Plans must use EPA's  reference methods  or EPA-approved  equivalent
 methods.
      2.   Calibration procedure  review -  Calibration  procedures
 used for all  measurement  methods should be reviewed,  revised if
 necessary,  documented,  and  included  in the  method writeup  just
 mentioned.   Document  control should also be established for these
 calibration  procedures  to inform  agency personnel of  any changes.
 As  an  agency  policy,  traceability of the  accuracy  of   working
 calibration  standards  should  be established  by  comparing these
 standards  to standards of higher accuracy whenever standards of
 higher  accuracy are available.
      3.   Internal  quality control procedures -  The procedures
 used  during  sampling  and  analysis to  detect,  correct,  and record
 out-of-control conditions  should  be defined and  documented.  Use
 of control charts is encouraged.
     4.   Audit .performance  -  Procedures should  be selected and
 implemented that will permit comparison of the performance of the •
measurement  system  (sampling and analysis) under routine opera-
tion versus an independent technique.   Commonly,  this independent
technique is either  a  performance  audit' or   the  use  of  a dual
measurement  system.    Results  from  these  audit  procedures  are
useful in detecting bias in the routine  measurement system.

-------
                                             Section No. 2.0.5
                                             Revision No. 0
                                             Date May 1, 1978
                                             Page 5 of 5

     5.   Interlaboratory testing - Each  agency and its contrac-
tors conducting  monitoring activities should  participate  in the
EPA quality assurance performance surveys.  Requests for partici-
pation should be made at the EPA regional office.
5.2.3  Develop Within Twelve Months -
     1.   Data  validation  procedures -  The   criteria  used  to
validate  air  monitoring  data  should be  documented,  and  the
routine tests or checks on the raw data should be defined.
     2.   Preventive  maintenance -   A schedule  for  preventive
maintenance should be prepared that identifies the required main-
tenance  tasks  and  frequencies.•.  A  procedure  for performing the
maintenance tasks should be prepared if none is available.
     3.   Review of training  needs  - Proper  training  of  staff
members  is  essential  for the  performance of  their  assigned job
responsibilities.   During  the  first 12  mo,   the  training  and
experience  of  all  staff  members should  be reviewed,  and  plans
should be made to obtain needed training.

-------

-------
Section 2.0.6   Chain-of-Custody Procedures for Ambient Air Samples
This section is up-to-date except that data logger printouts have now replaced
most  strip  chart recording systems for  ambient air monitoring.   Also, the
manual, wet chemical methods for SO2 and NO2 that use bubbler solutions, as
mentioned in Figures 6.1 and 6.2, are now rarely used because automated SO2
and NO2 air monitoring instruments are available.

-------

-------
                                              Section No.  2.0.6
                                              Revision No.  0
                                              Date July 1,  1979
                                              Page 1 of 11
 6.0  CHAIN-OF-CUSTODY PROCEDURES FOR AMBIENT AIR SAMPLES
      A quality  assurance  program associated with  the  collection
 of ambient  air  monitoring data must include an  effective  proce-
 dure for preserving the integrity of the data.   Ambient air test
 results and, in certain types  of tests,  the sample itself  may be
 essential elements in  proving the compliance status of  a  facil-
 ity;  that is, it may be necessary to introduce  the sample  or the
 test results  as evidence  in an  enforcement proceeding.   These
 will  not be  admitted  as evidence unless  it can be shown that they
 are representative of  the  conditions  that  existed at  the  time
 that  the  test was conducted.   Showing  this  requires  that  each
 step  in the  testing and analysis procedure be carefully monitored
 and documented.
      There are basically  four  elements  in the evidentiary  phase
 of an overall  quality.assurance  program:    •
      1.    Data  collection  - includes  testing,   preparation  and
 identification of the sample, strip  charts,  or -other data.
      2.    Sample  handling  -  includes  protection from  contamina-
 tion  and tampering during  transfer  between individuals  and  from
 the sampling site to the evidence locker.
      3.    Analysis  -  includes  storage of samples  prior to  and
 after  analysis as well  as data interpretation.
      4.    Preparation and  filing of  test report -  includes  evi-
 dentiary  requirements and retention of records."
 Failure  to include any one' of  these  elements  in the'collection
 and  analysis of ambient air monitoring  data may render the  re-
 sults  of the program inadmissible as  evidence,  or may seriously
 undermine  the credibility  of  any report  based on these  data.
 6.1  Sample Collection
     Ambient air  sampling  is primarily  concerned' with  the  atmo-
 spheric  concentrations  of  such  pollutants as  particulates,  SO
NO , CO, and photochemical  oxidants.

-------
                                              Section No. 2.0.6
                                              Revision No. 0
                                              Date July 1, 1979
                                              Page 2 of 11

     To establish the basic validity of such  ambient air monitor-
ing data, it must be shown that  (1) the proper methods were used,
(2) the  equipment was  accurately calibrated, and (3) the operat-
ing  technician  and  data analysts were qualified  and competent.
     The  data analyst  generally has  the primary responsibility
for determining  that the proposed  sampling method complies with
the  appropriate  testing regulations   and  that the  equipment is
accurately  sited.   The  technician is responsible for calibration
and operation of the monitors.   Each should be  able to support
and  justify the  test  methods  and  calibration procedures  used,
especially  in instances  where   it  is  necessary to  deviate from
accepted  practices.   For  example,  if  the only reasonable, test
site has  a  less  than ideal location, the network analyst must be
competent to  make a judgment based  upon  training  and experience
as  to  whether a -representative  sample  can  .be. obtained  at  the
site.   This  .determination,  should   be  recorded  and  included
in  the  program's protocol.  An  after-the-fact  site.analysis' may
suffice in  many  instances,  but good quality  assurance techniques
dictate  that  this  analysis  be  made  prior to  spending  the many
man-hours required  to  collect the, data.   Similarly,  the techni-
cian must be  able  to  confidently assert  that  the  equipment was
accurately  calibrated  using correct and  established calibration
methods.
6.1.1  Preparation  - Prior to  the  implementation of  a  sampling
and analysis  program,  a variety  of  sampling and  analysis equip-
ment must be  calibrated.  All data  and calculations involved in
these calibration activities should  be recorded in a calibration
log book.   It is suggested  that this log  be arranged so that" a
separate  section is designated  for each apparatus  and  sampler
used in the program.
     In  some  cases,  reagents  are  prepared  prior  to  sampling.
Some of  these reagents  will be  used to calibrate  the equipment,
while others  will become an integral part  of the  sample itself.
In any  case,  their integrity must  be carefully maintained from

-------
                                              Section No.  2.0.6
                                              Revision No.  0
                                              Date July 1,  1979
                                              Page 3 of 11

 preparation through analysis.  If there  are  any doubts about the
 method .by which the reagents  for  a  particular test were prepared
 or about  the competence of  the  laboratory technician preparing
 these items, the credibility of  the ambient air samples  and the
 test  results will be diminished.  It  is  essential  that a  careful
 record be kept  listing  the  dates  the reagents were  prepared,  by
 whom,  and  their locations  at all  times" from preparation  until
 actual use.   Prior  to  the test,  one  individual should be  given
 the responsibility  of -monitoring the handling  and  the use of the
 reagents.  Each  use  of the reagents  should be  recorded in  a  field
.notebook.                 .      .          ,
      Similarly,  filters  must be  selected and  prepared prior  to
 sampling.   These should be inspected to  assure  that  there are  no
 pinholes,  tears, creases,  or other  flaws  which may  affect  the
 collection  efficiency of the filter.   Each  step  in  the  filter
 equilibration, weighing,  and handling procedures should be  care-
 fully  recorded  by the/technician.to  assure  that the  ambient  air
 sample  obtained  with each  filter' adequately represents existing
 conditions.
 6-l-2   Identification  - Care must be taken to properly mark  all
 samples  and monitoring device readings to ensure  positive  identi-
 fication throughout  the  test and analysis procedures.  The  rules
 of evidence used, in  legal proceedings require that procedures  for
 identification of samples  used  in  analyses  form  the. basis  for'
 future evidence.   An.admission by the laboratory  analyst that he/
she cannot  be positive  whether  he/she analyzed  sample No.  6 or
sample  No.  9, for  example,  could destroy  the validity  of the
entire test report..  •
     Positive identification  also, must be provided for any fil-
ters used in the program.  If ink is used for marking,  it must be
indelible and unaffected by the gases and temperatures to which
it will  be  subjected.   Other methods  of identification  can be
used,   if  they provide a positive means of identification  and do
not impair the capacity of the filter to  function.

-------
                                              Section No.  2.0.6
                                              Revision No.  0
                                              Date  July  1,  1979
                                              Page  4 of  11

      Strip  charts from  automated  analyzers must also  be  clearly
 and  unambiguously  identified.   The  information  must  be placed
 upon  each strip chart  so as not  to  interfere  with  any of  the data
 on  the chart. If the  strip chart  is very long,  the information
 should be  pla'ced   at  periodic  intervals  on  the  chart.   The
 markings  should  be indelible  and  permanently affixed  to each
 strip  chart.
      Finally,  each  container  should have a unique identification
 to  preclude the  possibility  of  interchange.   Grease pencils may
 be  used for this purpose; a better method,  however,  is to affix
 an  adhesive-backed . label  to  the  container.   The number' of the
 container  should be subsequently  recorded on  the analysis data
 form.   Figure 6.1   shows  a' standardized  identification  sticker
 which  may  be used.  Additional  information  may be  added as re-
 quired, depending on the particular monitoring program.'
 6.2  Sample Handling
     If  actual •  samples  are  cpllected,  they  must be properly
 handled  to  ensure  that there is  no contamination and that the
 sample analyzed, is actually the sample taken  under the  conditions
 reported.   For  this reason, samples  should  be  kept  in a secure
 place  between the time  they, are collected  and  the time they are
 analyzed.   It is highly recommended  that  all samples  be secured
 until discarded.   These  security measures should be documented by
 a written record,  signed  by the handlers of the sample.
 6-2.1  Contamination and Tampering - To reduce the possibility of
 invalidating  the  results,  all  collected  samples  must be care-
 fully,  removed    from   the  monitoring  device  and  -placed  in
 sealed,  nonreactive  containers.   The  best  method  of  sealing
 depends on.  the container;  in  general, the  best  way  is  to simply
 use a  piece of tape to preclude accidental  opening of the con-.
 tainer  and  to  act  as  a  sufficient  safeguard  where  all other
 aspects of the chain-of-custody procedure are observed.   However,
when there  is any possibility  of temporary access to'the samples
by unauthorized personnel, the sample jars, containers,  or

-------
                                           .Section ,No.  2.0.6
                                           Revision No.  0
                                           Date July 1,  1979
                                           Page 5 of 11
Site name rjc/WZ ^fT
Site address  --#"33 18
<*T — &O/& • &/J- J3S~0
Date collected ^^_-7^ J~r>?/eT
Signature £krh-— • s&l-o-*—
U
Figure 6.1  Label for sample identification showing typical
            entries for SO2 bubbler solution.

-------
                                              Section No.  2.0.6
                                              Revision No.  0
                                              Date  July 1,  1979
                                              Page  6  of 11

 envelopes  should be sealed with a self-adhesive  sticker which has
 been  signed  and  numbered  by  the  operating  technician.  This
 sticker must adhere  firmly to ensure  that  it cannot  be  removed
 without destruction.   The  samples should  then be delivered to the
 laboratory for  analysis.  It is  recommended  that  this  be  done on
 the  same -day that the  sample  is  taken  from the monitor.   If this
 is  impractical,  all  the samples should  be  placed in  a carrying
 case  (preferably  locked)  uin  which •  they   are   protected  from
 breakage,  contamination, and loss.
      In transporting  samples  and other  monitoring  data,  it is
 important  that precautions be taken to .eliminate  the possibility
 of tampering,  accidental destruction,  and/or physical  and chemi-
 cal  action on the sample.  These practical  considerations must
 be dealt with on-a  case-by-case basis.
      The person  who has custody  of the samples, .strip  charts, or
 Other data must be  able to testify that no tampering occurred.
 Security must  be  continuous.    If  the  samples  are  put • in  a
 truck—lock  it.   After  delivery to  the  laboratory,  the  samples
 must be  kept in  a secured place.
      To  ensure that none of the sample is lost in  transport, mark
 all  liquid levels  on the  side  of  the  container with a grease
 pencil.   Thus,   any major  losses which  occur  will  be   readily
 ascertainable.
 6-2-2   Chain of  Custody -  If  the results of a  sampling  program
 are  to be  used  as  evidence,  a written record  must be available
 listing  the  location of the  data at all times.'  This chain-of-
 custody  record is  necessary to make  a prima  facie showing of the
 representativeness  of  the  sampling data. Without  it,  one cannot
be sure  that the sampling  data analyzed was the same as the data
proported  to  have  been taken  at a particular  time.  The  data
 should be handled only by persons associated in some way with the
 test  program.  A good  general  rule to  follow is "the fewer hands
the   better,"   even    though    a  properly   sealed   sample,
 for example,  may pass through a number  of hands without affecting
its integrity.

-------
                                             "Section No.  2.0.6
                                              Revision No. 0
                                              Date July 1, 1979
                                              Page 7 of 11

      Each person  handling the  samples or  strip charts must be
 able  to  state  from  whom  the item  was  received  and  to  whom
 delivered.   Recommended practice  is to have each  recipient  sign
 a chain-of-custody form  for the sampling data.  Figure  6.2  is  a
 form which may be  used  to establish the chain of  custody.   This
 form must- accompany the  samples or strip charts at  all times  from
 the  field  to ~ the  laboratory.   All persons  who  handle  the  data
 must sign the  form.
      When using the  U.S.  Postal  Service  to  transport  sampling
 data,  only certified  or  registered mail should be  used,  and  a
 return receipt .should be  requested.  The return receipt should be
 marked to  indicate  that the package  is to be delivered to  the
 addressee only.  The  addressee should be the specific person who
 is authorized  to receive  the  data.
     When using  the United Parcel Service,  commercial bus  lines,
 or similar, means of shipment, information describing  the  enclosed
 sampling  data  should be  placed  .on  the Bill  of  Lading.    The
 package  should  be marked  "Deliver to  Addressee  Only,"  and it
 should be addressed to  the specific person  authorized to receive
 the  data.
 6.3  Analysis  of the Sample
     For  ambient air-  samples to  provide useful  information . or
 evidence, laboratory analyses must meet the following four basic
 requirements:
     .1.   Equipment must be  frequently and  properly  calibrated
 and maintained.
     2.   Personnel must be qualified to make the analysis.
     3-   Analytical -procedures  must be  in accordance with   ac-
cepted practice.
     4-   Complete  and accurate records must be kept.
The  first three  requirements  are  similar  to those previously
discussed, and need no further  elaboration.  Proper records  may
consist of a laboratory  notebook 'or summary sheets which contain
hourly  average concentrations from  strip chart readings.  Where

-------
                                              Section No.  2.0.6
                                              Revision No.  0
                                              Date  July 1,  1979
                                              Page  8  of 11
Plant
Sample
number
5!<6f5- ,
-23;??
J381










Number
of
container
1
1
1










Description
of samples
A;-\Jol •£•;/•{-££ ?W 2)/'io*J C-A#£r
SO*. 6a2>£j£/Z ' So/*T/etO /9/i/a/ d#-r# S/)££T
rf
3.3 J-9














Relinquished
by
~Jchn ~Do£
"70s"' 3~o*J€s
-rZ*i TtfK/rs
"7^/n .'TrfVfS














Received
by
'Tom 3~od£S
2)&M;e dfat
frs* C/ae.;?
37/»i C//w.£














Time
He 30
tftso
/ a a O
/ o o o














Date
//-Z/-1?
U>2t~78
//-it -7?
//-2i-T%














Reason for change
of custody
•fa A ' /rtA /QfA Ivs/S
/)s4/\ts/s a-fl -TS?
Ar/rt/vj/'s of S0*
/Qsfrt/jSJ's a-S /I/O*,
'.-. 	 ^ 	 . 	 1 	 fJC. 	 ^t 	













      Figure 6.2  Ambient air sample integrity form showing
                  typical entries.

-------
                                              Section No. 2.0.6
                                              Revision No. 0
                                             • Date July 1, 1979
                                              Page 9 of 11

 practical,  standard preprinted  forms  should be  used.  .Do not
 discard  these  records  since  it  is  possible  that  they will  be
 required to substantiate the final report in the future.
      It is  important to realize  that chain-of-custody procedures
 discussed earlier in this report do not stop with delivery of the
 sampling  data  for  analysis.    Monitoring  data  which  are  not
 immediately analyzed  should be  stored  in a •secure  location  to
 which only  authorized  persons have  access.  The analyst must  be
 prepared  to testify that  at all  times the  sampling  data  were
 either in his/her possession and view or in a secure place.   Once
 the  data  are   analyzed,  they should  be returned to  the  secured
 storage location and retained at least until the report has been
 finally accepted.
 6.4  Field Notes
      Manual recording  of data is sometimes  required  for  ambient
 air tests.  Standardized forms should be utilized to  ensure that
 all  necessary  information is obtained..  These  forms should  be
'designed to clearly identify the .process  tested,  the date and
 time, location  of  test station,  and  operating personnel.   These
 data may determine the' credibility of the  data and  should  no.t  be
 erased or  altered.  Any errors  should be crossed  out  with  a
 single line, and the correct value recorded above  the  crossed-out
 number.
      Do  not  discard the original  field  records  even  if.  they
 become soiled.   Copies  are not normally admissible as  evidence.
 For neatness,   the  field data may be  transcribed  or copied for
'incorporation in.a  final report,  but  the originals should be  kept
 on  file.   Since these  records may be  subpoenaed,  it is  important
 that all  field  notes be legible.
 6.5   The Report as  Evidence
      In  addition to samples  and  field  records, the report of the
 analysis  itself  may serve  as material  evidence.   Just  as the
procedures and  data  leading up to the  final  report are subject  to
 the  rules  of evidence,  so  is the  report itself.   Written docu-

-------
                                              Section  No.  2.0.6
                                              Revision No.  0
                                              Date  July  1,  1979
                                              Page  10  of 11

ments,  generally  speaking,  are considered  as  hearsay, and are not
admissible  as  evidence without  a  proper foundation.  A proper
foundation  consists  of introducing  testimony  from  all  persons
having  anything  to  do with the  major portions of  the test and
analysis.  Thus the  field operator, all persons  having  custody of
the  samples,  and  the  analyst would be  required to  lay the  founda-
tion for the  introduction of the  test  report  as  evidence.
      Legal rules  recognize  that a record  of  events is  the result
of  input  from many  per sons*'"who  have  no  reason to  lie and that
production of all these persons  as  witnesses is onerous.  These
rules  recognize  the  complexity  and  mobility of our  society and
are  relatively liberal.   Indeed in  many cases the  trial 'judge
will  require  the parties to stipulate as  to  the authenticity of
ambient  air   test   reports  during  the  pretrial   proceedings.
However, the  party against whom  the  report  is  offered still has
the  right  to cross-examine the  sampling  program participants if
the  party  has reasonable cause.   In  this area,  the trial judge
may exercise  discretion.
     The relaxed  attitude  toward reports  of  experiments made by
persons in the regular course  of activity greatly simplifies the
introduction  of  the report in  evidence.   Only  the' custodian of
the report need testify in most cases.
     To  ensure compliance  with  legal rules,   all  test  reports
should  be  filed  'in  a safe  place by  a custodian having this re-
sponsibility.  Although the field notes and calculations  are not
generally included in the  summary report,  these materials may be
required at a future date to bolster the acceptability  and credi-
bility .'of  the report  as evidence in  an  enforcement proceeding.
Therefore,   the  full  report  including  all  original'  notes  and
calculation sheets  should  be  kept in  the  file.   Signed receipts
for  all  samples,  strip  charts,   or  other 'data,  should  also  be
filed.
     The original of  a document  is  the best evidence; and a copy
is  not normally  admissible  as  evidence.   Microfilm,  snap-out

-------
                                             Section No. 2.0.6
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 11 of 11

carbon  copies,  and  similar  contemporary  business  methods  of
producing copies are acceptable in many jurisdictions if unavail-
ablility of the  original  is adequately explained and if the copy
was made in the ordinary course of business.
     In summary, although all original calculations and test data
need not be included  in the final report, they should.be kept in
the  agency's   files.   It  is  a  good  rule  to  file  all  reports
together in a secure place.   Keeping  these documents under lock
and  key will  ensure that the  author can  testify  at future court
hearings that the report has not been altered.

-------

-------
                           March 1985
                                                                       Section Z.O.9
 9.0   QUALITY ASSURANCE IN THE OPERATION OF AUTOMATED AND MANUAL
               MONITORING METHODS IN SLAMS AND PSD NETWORKS
  On May 10, 1979, the U.S. Environ-
mental Protection Agency, in re-
sponse to Section 319 of the Clean
Air Act amendments of 1977, pro-
mulgated regulations set forth in 40
CFR Part 581 specifying ambient
monitoring requirements for State
Implementation Plans  (SIP's). These
regulations establish a national mon-
itoring network and provide uniform
monitoring requirements for all State
monitoring networks.

  An Important part of these specific
requirements is the establishment by
monitoring organizations of quality
assurance programs covering all as-
pects of their ambient monitoring.
Appendix A2 of Part 58 describes  the
quality assurance requirements for
organizations responsible for State
and'local air monitoring stations
(SLAMS). Appendix B3 of Part 58 de-
scribes the quality assurance require-
ments for organizations responsible
for prevention of significant deterio-
ration (PSD) air monitoring. Specifi-
cally, Section 2 of Appendices A and
B requires that the affected monitor-
ing organizations develop and imple-
ment quality assurance programs
consisting of policies,  procedures,
specifications, standards, and docu-
mentation necessary to: (1) provide
data of adequate quality to meet
monitoring objectives  and (2) mini-
mize loss of air quality data due to
malfunctions or out-of-control condi-
tions.
  The purpose of the following is  to
provide guidance, recommendations,
and suggestions for the development
and implementation of suitable qual-
ity assurance programs, as required
by Appendices A and B, related to
the operation  of continuous auto-
mated analyzers and manual meth-
ods used in SLAMS and PSD moni-
toring networks. The guidance
provided  here is directed mainly to-
ward establishing  a minimum level
of quality assurance for SLAMS and
PSD monitoring. Additional, more ex-
tensive quality assurance is encour-
aged. Reference 4 and  the measure-
ment methods sections of this
Handbook (Volume II) provide addi-
tional quidance.
 9.1  Quality Assurance for
 Automated Analyzers

 9.1.1  Analyzer Selection and Accep-
 tance Tests—Except for the specific
 exceptions described in Appendix C5
 of Part 58, monitoring methods used
 for SLAMS monitoring must be refer-
 ence or equivalent methods desig-
 nated as such by the USEPA.6 Moni-
 toring methods selected for PSD
 monitoring of gaseous criteria pollu-
 tants are restricted to automated ref-
 erence or equivalent methods.7 Ap-
 proval of any Appendix C exceptions
 for PSD monitoring must be obtained
 from the permit-granting authority
 before PSD monitoring is initiated.
 Section 2.0.4 of this Handbook (Vol-
 ume II) provides more information
 on reference and equivalent analyz-
 ers. A list of designated reference
 and equivalent methods may be ob-
 tained from the quality assurance co-
 ordinator of any USEPA regional  of-
 fice or from the Environmental
 Monitoring Systems Laboratory,
 Quality Assurance Division (MD-77),
 Research Triangle Park, North Caro-
 lina 27711.
  Among  reference and equivalent
 methods, a variety of analyzer de-
 signs and features are available. For
 some pollutants, analyzers employ-
 ing different measurement principles
 are available. And some analyzer
 models provide a higher level of per-
 formance than others that may only
 meet the minimum performance
 specifications.  Accordingly, in select-
 ing a designated method for a partic-
 ular monitoring application, consider-'
 ation should be given to such
 aspects as the  suitability of the mea-
 surement principle, analyzer sensitiv-
 ity and susceptability to interferences
 that may be present at the monitor- '
 ing site, requirements for support
 gases or other equipment, reliability
 and maintenance requirements, ini-
tial as well as operating costs, fea-
tures such as internal or fully auto-
matic zero and span checking or
adjustment capability, etc. Refer-
ence 8 and a series of four analyzer
reports (References 9-12) may be
helpful in evaluating and selecting
automated analyzers.
   It is important that the purchase
 order for a new reference or equiva-
 lent analyzer specify designation by
 EPA and .document the required per-
 formance specifications, terms of the
 warranty, time limits for delivery and
 for acceptance testing, and what hap-
 pens in the event that the analyzer
 delivered fails short of the require-
 ments.8 Upon receiving the  new ana-
 lyzer, the user should carefully read
 the instruction or operating  manual
 provided by the manufacturer of the
 analyzer. The manufacturer's manual
 should contain information or in-
 structions concerning:
   1.  unpacking  and verifying that all
   component parts were delivered;
   2. checking  for damage during
   shipment;
   3.  checking for  loose fittings and
   electrical connections;
   4.  assembling the analyzer;
   5.  installing the analyzer;
   6.  calibrating the analyzer;
   7.  operating the  analyzer;
   8.  preventive maintenance  sched-
   ule and procedures;
   9.  trouble shooting;-
   10.  list of expendable parts.
   Following analyzer assembly, an
 initial  calibration should be per-
 formed to determine if the analyzer
 is operating properly. Analyzer per-
 formance characteristics such as re-
 sponse time, noise, short-term span
 and zero drift, and precision should
 be checked during the initial calibra-
 tion or measured by using abbrevi-
 ated forms of the test procedures
 provided in 40 CFR  Part  53.6 Accep-
 tance of the analyzer should be
 based on results from these perfor-
 mance tests.8 Once accepted, refer-
 ence and equivalent analyzers are •
 warranted by the manufacturer to
 operate 'within the required perfor-
 mance limit for one year.6

 9.1.2  Analyzer Calibration—Calibra-
 tion of an analyzer establishes the
 quantitative relationship between ac-
'tual pollutant, concentration input (in
 ppm, ppb, (ig/m3, etc.) and the ana-
 lyzer's response (chart recorder read-
 ing, output volts, digital output, etc.).
 This relationship is  used to convert
 subsequent analyzer response values

-------
                             Section 2.0.9
                                                                           March 1985
  to corresponding pollutant concen-
  trations. Since the response of most
  analyzers has a tendency to change
  somewhat with time (drift), the cali-
  bration must be updated (or the ana-
  lyzer's response must be adjusted)
  periodically to maintain a high de-
  gree of accuracy.
    Each analyzer should be calibrated
  as directed by the analyzer's opera-
  tion or instruction manual and in ac-
  cordance with the general guidance
  provided here. For reference meth-
  ods for CO, NO2, and O3, detailed
  calibration procedures may also be
  found in the  appropriate appendix to
  40 CFR Part 50. Additional calibration
  information is contained in Refer-
  ences 13, 14, 15, and 16.
    Calibrations should be carried out
  at the field monitoring site by allow-
  ing the analyzer to sample test
  atmospheres containing known pol-
  lutant concentrations. The analyzer to
  be calibrated should be in operation
  for at least several hours (preferably
  overnight) prior to the calibration so
  that it is fully warmed up and its op-
  eration has stabilized. During the
  calibration, the analyzer should be
  operating in its normal sampling
 mode, and it  should sample the test
 atmosphere through all  filters, scrub-
 bers, conditioners, and other compo-
 nents used during normal ambient
 sampling and through as much of
 the ambient air inlet system as is
 practicable.* All operational adjust-
 ments to the analyzer should be
 completed prior to the calibration
 (see section 9.1.2 (g)). Analyzers that
 will be used on more than one range
 or that have autoranging capability
 should be calibrated separately on
 each applicable range.
   Calibration documentation should
 be maintained with each analyzer
 and also in a central backup file. Doc-
 umentation should be readily avail-
 able for review and should include
 calibration data, calibration equation
 (and curve, if prepared), analyzer
 identification,  calibration date, ana-
 lyzer location, calibration standards
 used and their traceabilities, identifi-
 cation of calibration equipment used,
 and the person conducting the cali-
 bration.
•Deviations from this general rule may be ac-
 ceptable for some CO analyzers to reduce con-
 sumption of calibration gas or to accommo-
 date automatic calibration systems because
 CO is more stable than other gaseous criteria
 pollutants and the scale range is normally 100 '
 limes higher. However, such deviations should
 be used with caution to avoid possible calibra-
 tion errors or bias.
   (a)  Calibration Standards—In gen-
   eral, ambient monitoring instruments
   should be calibrated by allowing the
   instrument to sample and analyze
   test atmospheres of known concen-
   .trations of the appropriate pollutant
   in air. All such (non-zero) test con-
   centrations must be, or be derived
   from, local or working standards
   (e.g., cylinders of compressed gas or
   permeation devices) that are certified
   as traceable to an NBS primary
   standard. "Traceable"  is defined in
   40 CFR Parts 50 and 58 as mean-
   ing" ... that a local standard has been
   compared and certified, either di-
   rectly or via not more than one inter-
   mediate standard, to a primary
   standard such as a National Bureau
   of Standards Standard Reference Ma-
   terial (NBS SRM) or a USEPA/NBS-
   approved  Certified Reference Mate-
   rial (CRM)." Normally, the working
  standard should be certified directly
  to the SRM or CRM, with an interme-
  diate standard used only when nec-
  essary. Direct use of  a CRM as a
  working standard is acceptable, but
  direct use  of an NBS SRM as a work-
  ing standard is discouraged because
  of the limited supply and expense of
  SRM's. As a minimum, the certifica-
  tion procedure for a working stand-
  ard should (a) establish the concen-
  tration of the working standard
  relative to  the primary standard,
  (b) certify that the primary standard
  (and hence the working standard) is
  traceable to an NBS primary stand-
  ard, (c) include a test of the stability
 ' of the working standard over several
  days, and (d) specify a recertification
  interval for the working'Standard.
  Certification of the working standard
  may be established by either the
  supplier or the user of the standard.
  A recommended protocol for certify-
  ing'gaseous standards against an
  SRM or CRM is given in Section 2.0.7
  of this Handbook (Volume II). Also, a
  list of CRM sources is available from
  the Quality Assurance Division
  (MD-77), Environmental Monitoring
  Systems Laboratory, U.S. Environ-
  mental Protection Agency, Research
' Triangle Park, North Carolina 27711.
   Test concentrations of ozone must
  be traceable to a primary standard
 UV photometer as described in Ap-
 pendix D of 40 CFR Part 50. Refer-
 ence 17 describes procedures for cer-
 tifying transfer standards for ozone
 against UV  primary standards.
   Test concentrations at zero concen-
tration are considered valid  stand-
ards. Although zero standards are
not required to be traceable to a pri-
  mary standard, care should be exer-
  cised to ensure that zero standards
  are indeed adequately free of all sub-
  stances likely to cause a detectable
  response on the analyzer. Periodi-
  cally, several different and indepen-
  dent sources of zero standards
  should be compared.  The one that
  yields the lowest response can usu-
  ally (but not always!)  be assumed to
  be the best zero standard. If several
  independent zero standards produce
  exactly the same response, it is likely
.  that all the standards are adequate.
    The accuracy of flow measure-
  ments is critically important in many
  calibration procedures. Flow or vol-
  ume measuring instruments should
  be calibrated and certified at appro-
  priate intervals (usually 3 to 6
  months) against NBS or other au-
  thoritative standards such as a trace-
  able bubble flow meter or gas meter.
  Calibration procedures for some
  types of flow and volume meters
  may be found in section 2.1.2.1
  (Vol. II) and section  3.5.2  (Vol. Ill) of
  this Handbook.
   Documentation of all calibrations
  of instruments and  certification of
  standards should be maintained,
  showing calibration date, calibration
  procedure used, calibration data or
 curve, name of person  conducting
 the calibration, and  the date for the
 next calibration.

 (b)   Multi-point Calibrations—
 Multi-point calibrations consist of
 three or more test concentrations, in-
 cluding zero concentration, a concen-
 tration between 80% and 90% of the
 full scale range of the analyzer under
 calibration, and one  or  more
 intermediate concentrations spaced
 approximately equally over the scale  •
 range. Multi-point calibrations are
 used to establish or  verify the linear-
 ity of analyzers upon initial installa- •
 tion and after major  repair.  Most
 modern analyzers have a linear or
 very nearly linear  response with con-
 centration. If a non-linear analyzer is
 being calibrated, additional calibra-
tion points should be included to ad-
equately define the calibration rela-
tionship, which should be a smooth
curve. Multi-point  calibrations are
likely to be more accurate than two-
point calibrations because of the av-
eraging effect of the  multiple points
and because an error in the genera-
tion of a test concentration (or in
recording the analyzer's response) is
more likely to be noticed as a point
that is inconsistent with the others.
For this reason, calibration  points
should be plotted or  evaluated statis-

-------
                              March 1985
                                                                           Section 2.0.9
  tically as they are obtained so that
  any deviant points can be investi-
  gated or repeated immediately.
    Most analyzers have zero and span
  adjustment controls, which should be
  adjusted based on the zero and high-
  est test concentrations, respectively,
  to provide the desired scale range
  within the analyzer's specifications
  (see section 9.1.2.e). (Note—for ana-
  lyzers in routine operation, unad-
  justed ("as is") analyzer zero and
  span response readings should be
  obtained prior to making any zero or
  span adjustments—see the discus-
  sion of unadjusted readings under
  "Level 1 zero and span calibration".)
  NO/NO2/NOX analyzers may not have
 •individual zero and span controls for
  each channel; the analyzer's opera-
  tion/instruction manual should be
  consulted for the proper zero and
  span adjustment procedure. Zero and
  span controls often interact with
  each other, so the adjustments may
  have to be repeated several times to
  obtain the desired final adjustments.
    After the zero and span adjust-
  ments have been completed and the
  analyzer has.been allowed to stabi-
  lize on the new zero and span set-
  tings, all calibration test concentra-
.  tions should.be introduced into the
  analyzer for the final, calibration. The
• final, post-adjusted analyzer re-
  sponse readings should be obtained
  from the same device (chart
  recorder, data acquisition system,
  etc..) that will be used for subsequent
  ambient measurements.
    The analyzer readings are plotted
  against the respective test concentra-
  tions, and the best linear (or non-
  linear if appropriate) .curve to fit the
  points'is determined. Ideally, least
  squares  regression analysis (with an
  appropriate transformation of the
  data for  non-linear analyzers) should
  be used  to determine the slope and
  intercept for the best fit calibration
  line of the form, y = mx + a, where y
  represents the analyzer response, x
  represents the pollutant concentra-
  tion,  m is the slope, and a is the
  x-axis intercept of the best fit calibra-
  tion line. When this calibration rela-
  tionship is subsequently used to
  compute concentration measure-
  ments (x) from analyzer response
  readings (y),  the formula is trans-
  posed to the  form, x = (y - a)/m. If
  the calibration points show very little
  deviation from the regression line
  (i.e., very little scatter), it may be ac-
'  ceptable (i.e., the error will be negli-
  gible) to do the regression "back-
•  wards," letting x represent the ana-
  lyzer  response and y represent the
 concentration. In this case, the cali-
 bration relationship, y = mx + a, is
 used directly, without transposition,
 to calculate concentration measure-
 ments (y) from analyzer response (x).
  As a quality control check on cali-
 brations, the standard error or corre-
 lation coefficient can  be calculated
 along with the regression calcula-
 tions. A control chart of the standard
 error or correlation coefficient could
 then be maintained to monitor the
 degree of scatter in the  calibration
 points and, if desired, limits of ac-
 ceptability could  be established.

 (e)  Level  1 Zero and Span Calibra-
 tion—A level 1 zero and span  calibra-
 tion is a simplified, two-point ana-
 lyzer calibration  used when analyzer
 linearity does not need to be checked
 or verified. (Sometimes  when  no ad-
 justments are made to the analyzer,
 the level 1 calibration may be  called
 a zero/span check, in which case it
 must not be confused with a level 2
 zero/span check  (see (d)). Since most
 analyzers  have a  reliably linear or
 near-linear output response with con-
 centration, they can be adequately
 calibrated with only two concentra-
 tion standards (two-point calibration).
 Furthermore, one of the standards
 may be zero concentration, which is
 relatively easily obtained and need
 not be certified. Hence, only one cer-
 tifed concentration standard is
 needed for the two-point (level 1)
 zero and span calibration. Although
 lacking the advantages of the multi-
 point calibration, the two-point zero
 and span calibration—because of its
 simplicity—can be (and  should be)
 carried out much more frequently.
Also, two-point calibrations are easily
 automated. Frequent checks or up-
 dating of the calibration  relationship
with a 2-point zero and span calibra-
tion improves the quality of the mon-
itoring data by helping to keep the
calibration relationship more closely
matched to any changes (drift) in the
analyzer response.
  As with  any calibration, the ana-
lyzer should be operating in its nor-
mal sampling mode, and generally
the  test concentrations should  pass
through as much of the  inlet and
sample conditioning system as is
practicable.  For NO2, SO2, and  partic-
ularly for QZ, wet or dirty inlet  lines
 and paniculate filters  can cause
changes in the pollutant concentra-
tion. Efforts should be made, at least
periodically, to introduce the span
calibration concentration into the
 sampling system as close to the out-
door sample inlet point as possible.
 The calibration response under these
 conditions can then be compared to
 the response when the span concen-
 tration is introduced at the analyzer,
 downstream of the sample inlet com-
 ponents, as a check of the entire
 sample inlet system.
   Some CO analyzers may be tempo-
 rarily operated at reduced vent or
 purge flows, or the test atmosphere
 may  enter the analyzer at a point
 other than the normal sample inlet,
 provided that such a deviation from
 the normal sample mode is permit-
 ted by the analyzer's operation or in-
 struction manual and the analyzer's
 response is not likely to be altered by
 the deviation. Any such operational
 modifications should be used with
 caution, and the lack of effect should
 be verified by comparing test calibra-
' tions made before and after the mod-
 ification.
   The standards used' for a level 1
 zero  and  span calibration must be
 certified traceable, as described pre-
 viously under "calibration stand-
 ards." The span standard should be
 a concentration between about 70%
 and 90%  of the analyzer's full scale
 measurement range.
  -Adjustments to the analyzer may
 be made during the zero and span
 calibration. However; it is strongly
 recommended that unadjusted (i.e.,
 "as is") analyzer response readings
 be obtained before any adjustments
 are made to the analyzer. As de-
 scribed later, these unadjusted zero
 and span readings provide valuable
 information for (1) confirming the
 validity of (or invalidating) the mea-
 surements obtained immediately pre-
 ceding the calibration, (2) monitoring
 the analyzer's calibration drift and
 (3) determining the frequency of re-
 calibration. Accordingly, the follow-
 ing procedure for a zero and span
 calibration is recommended:
    1.  Disconnect the analyzer's inlet
    from the ambient intake and con-
    nect it to a calibration system.
    Leave  the analyzer in its normal
    sampling mode, and make no
    other ajustments to the analyzer
    (except as mentioned previously
    for some CO analyzers).

   2. Sample and measure the span
   test concentration and record the
   unadjusted, stable ("as is") span
   response reading (S'). NOTE: All
   analyzer response readings should
   be  recorded in the analyzer's nor-
   mal output units, e.g., millivolts,
   percent of scale, etc. (the same
   units used for the calibration
  curve).  If these units are concentra-

-------
                             Section 2.0.9
                                                                           March 1985
    tion units they should be identified
    as "indicated" or "uncorrected" to
    differentiate them from the
    "actual" concentration units that
    are used for reporting actual ambi-
    .ent concentration measurements.
    3. Sample and measure the zero
    test concentration standard and
    record the unadjusted, stable zero
    reading  (2').
    4. Perform any needed analyzer
    adjustments (flow, pressure, etc.)
    or analyzer maintenance.
    5. If adjustment of the zero is
    needed (see subsections (e) and
    (f)) or if any adjustments have
    been made to the analyzer, adjust
    the zero to the desired zero read-
    ing. Offsetting  the zero reading
    (e.g., to 5% of scale) may help to
    observe  any negative zero drift that
    may occur. Record the adjusted,
    stable zero reading (Z). If no zero
    adjustment is made, Z= Z.'
    6. Sample and measure the span
    test concentration. If span adjust-
    ment is'needed (see  subsections
    (e) and {f}), adjust the span re-
    sponse to the desired value, allow-
    ing for any zero offset used in the
    previous step. Record the final ad-
   justed, stable span reading (S). If
   no adjustment is made, S = S.
   7.  If any  adjustments made to the
   zero, span, or other parameters or
   if analyzer maintenance was car-
   ried but, allow the analyzer to
   restafalize at the new  settings, then
   recheck the zero and  span readings
   and record new values for Z and S,
   if necessary.
   If the calibration is updated for
 each zero/span calibration (see sec-
 tion 9.1.3), the new calibration rela-
 tionship should be plotted using the
 Z and S readings,  or the intercept
 and slope should be determined as
 follows:
    intercept» Z
       slope<
                    S-Z
              span concentration
Id) Level 2 Zero and Span Check—A
level 2 zero and span check is an
"unofficial" check of an analyzer's re-
sponse. It may include dynamic
checks made with uncertified test
concentrations, artificial stimulation
of the analyzer's detector, electronic
or other types of checks of a portion
of the analyzer, etc.
  Level 2 zero and span checks are
not to be used as a basis  for analyzer
zero or span adjustments, calibration
updates, or adjustment of ambient
   data. They are intended as quick,
   convenient checks to be used be-
   tween zero and span calibrations to
   check for possible analyzer malfunc-
   tion or calibration drift. Whenever a
   level 2 zero and span check indicates
   a possible calibration problem, a
   level 1 zero and span (or multipoint)
   calibration  should be carded out be-
  fore any corrective action is taken.
    If a level  2 zero and span check is
  to be used  in the  quality control pro-
  gram, a" "reference response" for the
  check should be obtained immedi-
  ately following a zero and span (or
  multipoint)  calibration while the arfa-
  lyzer's calibration is accurately
  known. Subsequent level 2 check re-
  sponses should then be compared to
  the most recent reference response
  to determine if a  change in response
  has occurred. For automatic level 2
  zero  and span checks, the first sched-
  uled  check following the calibration
  should be used for the reference re-
  sponse. It should be kept in mind
  that any level 2 check that involves
  only part of the analyzer's system
  cannot provide information about the
  portions of the system not checked
  and therefore cannot be used as a
  verification of the overall analyzer
  calibration.

  (e)   Physical Zero and Span Adjust-
  ments—Almost ail ambient monitor-
  ing instruments have physical means
  by which to  make zero and span ad-
 justments. These adjustments are
  used to obtain the desired nominal
 scale  range (within the instruments'
 specifications), to  provide convenient
 (nominal) scale units, and to periodi-
 cally adjust the instruments' re-
 sponse to correct for calibration drift.
 Note: NO/N02/NOX analyzers may
 not have individual zero and span
 controls for each channel. If that is
 the case, the zero and span controls
 must be adjusted only under the con-
 ditions specified in the' calibration
 procedure provided in the analyzer's
 operation/instruction manual.
   Precise adjustment of the zero and
 span controls may not be possible
 because of (1) limited resolution of
 the controls,  (2) interaction between
 the zero and  span controls, and
 (3) possible delayed reaction to ad-
 justment or a substantial  stabilization
 period after adjustments are made.
 Precise adjustments may not be nec-
 essary, however, because calibration
 of the  analyzer following zero and
span adjustments will define the pre-
cise response characteristic (calibra-
tion curve). Accordingly, zero and
span adjustments must always be
  followed by a calibration. Allow suffi-
  cient time between the adjustments
  and the calibration for the analyzer to
  fully stabilize. This stabilization time
  may be substantial for some analyz-
  ers. Also, obtain unadjusted re-
  sponse readings before Adjustments
  are made, as described in the previ-
  ous section on level 1 zero  arid span
  calibration.
    Zero and span adjustments do not
  necessarily need to be made at each
  calibration. In fact, where only rela-
  tively small adjustments would be
 . made, it is probably more accurate
  not to make the adjustments because
  of the difficulty of making precise ad-
  justments mentioned earlier. An ap-
  propriate question, then, is  how
  much zero or span drift can be al-
  lowed before  a physical zero or span
  adjustment should be  made to an an-
  alyzer?
    Ideally, all ambient measurements
  obtained from an analyzer should be
  calculated or adjusted  on the basis of
  the most recent (zero and span or
.  multipoint) calibration  or on the
  basis of both the  previous and sub-
 sequent calibrations (see section
 9.1.3 on Data Processing). In this
 case, considerable drift (i.e., devia-
 tion from an original or nominal re-
 sponse curve) can be allowed before
 physical adjustments must be made
 because the calibration curve used to
 calculate the ambient measurements
 is  kept in close agreement with the
 actual analyzer response. The chief
 limitations are the amount of change
 in  the effective scale range of the  an-
 alyzer that can be tolerated and pos-
 sible loss of linearity in the analyzer's
 response due to excessive deviation
 from the design range.  Cumulative
 drifts of up to 20% or 25% of full
 scale from the original  or nominal
 zero and span values may not be
 unreasonable,  subject to the
 limitations mentioned above. '
  In situations  where it  is not possi-
 ble to update the calibration  curve
 used to calculate the ambient read-
 ings after each zero and span calibra-
 tion, then the ambient readings must
 be  calculated from the most recent
 multipoint calibration curve or from a
fixed nominal or "universal"  calibra-
tion curve (section 9.1.3). In this case,
the zero and span calibrations serve
only to measure or monitor the devi-
ation  (drift error) between the actual
analyzer response curve and the cali-
bration curve used to calculate the
ambient measurements. Since this
error must be kept small, physical
zero and  span adjustments  are much
more  critical and should be made be-

-------
                             March 1986
                                                                          Section 2.0.9
 fore the error becomes large. More
 information on drift limits and deter-
 mining when physical zero and span
 adjustments are needed is contained
 in the next section on frequency of
 calibration. See also'Figure 9-2.

 (f)  Frequency of Calibration and An-
 alyzer Adjustment—As previously in-
 dicated, a multipoint calibration
 should be carried out on  new analyz-
 ers  or after major repairs to establish
 analyzer linearity. It is also appropri-
 ate  to carry out a  multipoint calibra-
 tion on each analyzer in routine oper-
 ation at least twice per year to
 reverify linearity, although an annual
 multipoint audit may serve in lieu of
 one of these. (Nonlinear analyzers
 may, of course, require more fre-
 quent multipoint calibration if they
 cannot be calibrated adequately with
 2-point calibrations.)
   The calibrations referred to bglp_w
 would normally be 2-point zero and
 span (level 1) calibrations; however,
 a multi-point calibration can always
 substitute for a 2-point calibration.
 bration.
   An analyzer should be calibrated
 (or recalibrated):
   1. upon initial installation;
   2. following  physical relocation;
   3. after any repairs or service that
   might affect its calibration;
   4. following an interruption in oper-
   ation of more than a few days; and
   5. upon any indication  of analyzer
   malfunction or change in calibra-
   tion.
 In addition, analyzers in routine oper-
 ation should be recalibrated periodi-
 cally to maintain close agreement be-
 tween  the calibration relationship
 used to convert analyzer responses
 to concentration measurements and
 the actual response of the analyzer.
 The  frequency of this routine  peri-
 odic recalibration is a matter of judg-
 ment and is a tradeoff among several
 considerations, including:  the inher-
 ent stability of the  analyzer under the
 prevailing conditions of temperature,
 pressure, line voltage, etc. at the
 monitoring site; the cost and incon-
 venience of carrying out the calibra-
 tions; the quality of the ambient
 measurements needed; the number
 of ambient measurements lost dur-
 ing the calibrations; and the risk of
 collecting invalid data because of a
 malfunction or response problem
with the analyzer that wouldn't be
 discovered until a calibration is car-
 ried  out.
  When a new monitoring instru-
ment is first installed, level 1 zero
and span calibrations should be very
 frequent—rperhaps daily or 3 times
 per week—because little or no infor-
 mation is available on the drift per-
. formance of the analyzer. (Informa-
 tion on another unit of the same
 model analyzer may be useful; how-
 ever, individual units of the same
 model may perform quite different-
 ly.) After enough information on the
 drift performance of the analyzer has
 been accumulated, the calibration
 frequency can be adjusted to provide
 a suitable compromise among the
 various considerations mentioned
 above. However, prudence suggests
 that the  calibration frequency should
 not be less than every two weeks. If
 a biweekly frequency is selected and
 the level 1 zero/span calibration is
 carried out on the same day as the
 one-point precision check required in
 Subsection 3 of Appendices A and B
 of Part 58, the precision check must
 be done first.
   To facilitate the process of deter-
 mining calibration frequency, it is
 strongly recommended that control
 charts be used to monitor the zero
 and span drift performance of each
 analyzer. Control charts can be con-
 structed  in different ways, but the
 important points are to visually rep-
 resent and statistically monitor zero
 and span drift, and to be alerted if .
 the drift becomes excessive so that
 corrective action can be taken. Exam-
 ples of simple zero and span control
 charts are shown in Figure 9-1. Such
 control charts make important use of
the unadjusted zero and span re-
sponse readings mentioned in sec-
tion 9.1.2(c).
   In the zero drift chart of Figure 9-1,
cumulative zero drift is shown by
plotting the zero deviation in ppb for
each zero/span calibration relative to
a nominal calibration curve (inter-
cept = 0 scale percent, slope = 200
scale percent per ppm for a nominal
scale range of 0.5 ppm). This zero
deviation may be calculated as fol-
lows:
Z'-
  m0
               ;x 1000 ppb/ppm
where
   Dz = zero deviation from the refer-
       ence calibration (e.g., nomi-
       nal or original calibration),
       ppb;
   Z' = unadjusted zero reading, e.g.,
       scale  percent;
   I0 = intercept of reference calibra-
       tion, e.g., scale percent;
  m0 = slope of reference calibration,
       e.g., scale percent/ppm.
 Similarly, cumulative span drift may
 be shown by plotting the percent de-
 viation in the slope of the calibration
 curve relative to the reference cali-
 bration. This percent deviation in the
 span slope may be calculated as fol-
 lows:
                                  Ds

                   x 100 percent
 where
    Ds = span deviation from reference
        calibration, percent;
   m0 = slope of reference calibration,
        e.g., scale percent/ppm;
   mc = current analyzer  calibration
                S' — Z'
        slope = — g — , e.g., scale
        percent/ppm;
    S' = unadjusted span reading, e.g.,
        scale percent;
    Z' = unadjusted zero  reading, e.g.,
        scale percent;
    C = span concentration.

   Where physical zero or span ad-
 justments have been made to the an-
 alyzer (marked by diamonds along
 the horizontal axes in Figure 9-1),
 both the unadjusted (Z'f- S') and the
 adjusted readings (Z, S) are plotted
 (substitute Z for Z'  and S for S' in
 the- formulas). The connecting line
 stops at the unadjusted reading,
 makes  a vertical transition represen-
 tative of the physical adjustment,
 then  continues from the adjusted
 reading.
   The charts in  Figure 9-1 cover a pe-
 riod of 150 days, with zero/span cali-
 bration every 2 or 3 days  (2.7 days
 on the average). Practical adjustment
 limits were set at ±15 ppb for zero
 and ±7% for span,  (shown as broken
 lines in Figure 9-1), although most of
 the span adjustments and all of the
 zero adjustments were made before
•these limits were reached. These lim-
 its could have been set wider be-
 cause the calibration slope and inter-
 cept used to calculate the ambient
 readings were updated at each  zero/
 span  calibration. Narrower limits
 may be needed  if the calibration
 curve used to calculate the ambient
 data is not updated at each zero/span
 calibration.
  The total net cumulative zero drift
 over the entire 150 day period (ignor-
 ing zero adjustments) was  +.9 ppb,
 indicating that the analyzer's zero
stability was good. Total net cumula-
tive span drift (ignoring span adjust-
ments) was +15.4%, indicating that
the analyzer should be watched
closely for continued positive span
drift. Most of the individual zero and

-------
                             Section 2.0.9
                                                                            March 1985
Q

I
       +30

       +20

       +ro

       +0

       •10

       •20
      -30
                                                                                      Tot. Nat Zero Drift, ppb  +.9
                                                                                      Number of Drift Periods: 55
                                                                                      Ave Drift Period, days:   2.7
                                                                                      Ave[Drifty Period, ppb:  1.5
                                                                                      Std Dev. Zero Drift, ppb:  2.4
              1 - 1 - 1 - 1
                                            1   •
                                                        _1	1_
                                                                                             6
         30
                    45
                           60 '
                                        75
                                                  SO
                                                         JOS

                                                     Day of Year
                                                                     120
                                                                               135
                                                                                         150
                                                                                                   165
                                                                                                             180
                                                                                       Total Net Span Drift, %: +1S.4
                                                                                       Number of Drift Periods: 55
                                                                                       A ve Drift Period, days:    2.7
                                                                                       Ave [DriftyPeriod, %:    1.8
                                                                                       Std Dev. Span Drift. %:   2.3
                                                           105       120

                                                       Day of Year
                                                                            135
                                                                                       150
                                                                                                165
                                                                                                          180
 Ftgurs 9-1.    Examples of simple zero and span control charts.
 span drifts {i.e., the net change from
'one zero/span calibration to the next)
 were small. The average of the abso-
 lute values of these individual  zero
 drifts (ignoring zero adjustments)
 was 1.5 ppb, and the average of  the
 absolute values of the individual
 span drifts (ignoring span adjust-
 ments) was 1.8 percent. In view of
 these relatively low values, the fre-
 quency of zero/span calibrations
 could be reduced, say to twice a
week or every 4 days, particularly if
 level 2 zero/span checks were used
between the level 1 zero/span cali-  .
brations. However, such reduced  cali-
bration frequency would tend to  in-
crease the average error between the
actual analyzer response and the  cali-
bration curve used  to calculate the
ambient measurements. Reduced cal-
ibration frequency would also in-
crease the risk of collecting  invalid
data because of potentially increased
delay in discovering a malfunction or
serious response change. If either of
the average zero or average span
drift is large, more frequent zero/
                                    span calibration should be consid-
                                    ered.
                                      A final pair of statistics that should
                                    be calculated is the standard devia-
                                    tions of the individual zero and span
                                    drifts, respectively (again, ignoring
                                    zero and span adjustments). These
                                    values (2.4 ppb and 2.3%, respec-
                                    tively, for the charts shown in Fig-
                                    ure 9-1) provide a measure of-the
                                    typical drift performance of the ana-
                                    lyzer. A band equal to ±3 standard
                                    deviations can be established to rep-
                                    resent "normal" performance of the
                                    analyzer. Such a band is represented
                                    on the charts of Figure 9-1 by the
                                    l-bands at the right edge of the
                                    charts. Any excursion outside of
                                    these bands is an indication of a pos-
                                    sible performance problem that may
                                    need corrective action or additional
                                    scrutiny.
                                      In continual monitoring, the total
                                    cumulative  drift, average of the abso-
                                    lute values  of the individual drifts,
                                    and the standard deviation of the in-
                                    dividual drifts should be calculated
                                    on a running basis over  the last 100
 or so days. Figure 9-2 summarizes
 some of the ranges and control chart
 limits discussed previously. These
 limits are suggested,'but they could
 be modified somewhat at the discre-
 tion of the monitoring agency. There
 are also other ways.tqjonstruct con-
 trol charts. Appendices J and H of
 Volume I of this Handbook provide
 additional  information on the calcula-
 tion  of standard deviations and on
 the construction and interpretation of
 control charts.

 (g)  Automatic Self-Adjusting Ana-'
 lyzer—Some air monitoring analyz-
 ers are capable of periodically carry-
 ing out automatic zero and span
 calibrations and making their own
 zero and span self adjustments to
 predetermined readings. How should
 such automatic zero/span calibrations
 be treated? If the automatic zero/
 span calibration meets all the re-
 quirements discussed previously for
 level 1 zero and span calibrations
 (i.e.,  traceable standards that pass
through the sample inlet and sample

-------
                            March 1985
                                                                          Section 2.O.9
                      Calibration updated at each zero/span
          Zero
          Drift

       +20 to 30 ppb
       (2 to 3 ppm CO)
           +3 stddev



           +1 stddev

                  0

           -1 stddev


           -3 std dev
      -20 to -30 ppb
      (-2 to -3 ppm CO)
and recalibrate
Analyzer adj
option
Normal analyzer
range
\
'ustment
aT

Analyzer adjustment
not recommended

Analyzer adj

ustment
optional
I
Adjust analyzer
and recalibrate
    Span
    Drift

+20% to 25%
+3 std dev



+J stddev

 0

•1 stddev


•3 std dev
-20% to -25%
                      Fixed calibration used to calculate data
zero
Drift
+ 10 to 15 ppb
11 to 1.5 ppm CO)
+3 std dev
+ std dev
0 —
-/ stddev
-3 std dev
-JO to -15 ppb
(-1 to -1.5 ppm CO)
Invalidate data; adjust ,
and recalibrate analyzer
I
Adjust and
recalibrate analyzer
Normal analyzer J
1 range
Adjustment
optional
Analyzer adjustment
not recommended

Adjustment
optional
1
Adjust and
recalibrate analyzer
1
Invalidate data; adjust
and recalibrate analyzer
Span
Drift
+15%
+3 std dev
+1. std dev
— 0
-1 stddev
-3 std dev
•15%
Figure 9-2.    Suggested zero and span drift limits when the calibration used to calculate
             measurements is updated at each zero/span calibration (upper) and when a
             fixed calibration is used to calculate measurements (lower).
conditioning system) and both the
adjusted and unadjusted zero and
span response readings can be ob-
tained from the data recording de-
vice, then the calibration may be
treated as a valid zero/span calibra-
tion as discussed in this section. If
the automatic calibrations do  not
qualify as level 1 calibrations  (be-
cause the final zero and span  read-
ings cannot be read from the  strip
chart for example), then the analyzer
must receive manual zero/span cali-
brations as if it had no automatic ca-
pabilities. In this case, the automatic
zero and span adjustments should be
ignored, except that manual calibra-
tions should be separated in time as
much  as possible from the occur-
rence of the automatic calibrations
for maximal benefit. It may some-
times happen that automatic and
manual calibrations interact, produc-
ing a detrimental effect on the moni-
toring data. If so, the automatic cali-
brations should be discontinued  or
adjusted to avoid continuation of the
conflict.
(h)  Level 1 Zero and  Span Calibra-
tion Documentation—All Level 1  zero
or span calibrations should be docu-"
mented in a chronological format.
Documentation should include ana-
lyzer identification, date, standard-
used and its traceability, equipment
used, the individual conducting the
calibration, the unadjusted zero and
span responses, and the adjusted
zero and span responses. Again,
quality control charts are an excellent
form of documentation to graphically
record and track calibration results.
Level 1 zero and span  documentation
should be maintained  both in  a cen-
tral file and at the monitoring  site.
HI  Use of Computers for Control
Chart Plotting and Warning of Out-of-
Control Conditions—With .the wide
range of economical computers now
available, consideration should be
given to a computer system that can
process and output the information  -
in a timely fashion. Such a computer
system should be able to:
  1.  Compute calibration equations
  2.  Compute measures of linearity of
  calibrations (e.g., standard error or
  correlation  coefficient)
  3.  Plot calibration curves
  4.  Compute zero/span drift results
  5.  Plot zero/span drift data
  6.  Compute precision and accuracy
  results
  7.  Compute control chart limits
  8.  Plot control charts
  9.  Automatically flag out-of-control
  results

-------
                             Section 2.0.9
                                                                          March 1985
    10. Maintain and retrieve calibra-
    tion and performance records.

  9.1.3  Data Processing

  M   Calculation of Ambient Mea-
  surements—As noted previously, an
  analyzer's response calibration curve
  relates the analyzer response to ac-
  tual concentration units of measure,
  and the response of most analvzers
  tends to change (drift) unpredictably
  with passing time. These two condi-
  tions must be addressed in the
  mechanism that is used to process
  the raw analyzer readings into final
  concentration measurements. Four
  practical methods are described be-
  low. They are listed in order of pref-
  erence, with the first one  being the
  most likely to minimize errors caused
  by differences between the actual an-
  alyzer response and the response
  curve used to calculate the measure-
  ments. As would be expected, the
  order also reflects decreasing com-
  plexity and decreasing difficulty of
  implementation. The first 3 methods
  are best implemented with automatic
  data processing systems because of
  the number of calculations required.
  Methods 3 and, 4 could be used on a
  manual basis and are more labor in-
  tensive because of the need for more
  frequent and precise physical adjust-
  ment of analyzer zero and span  con-
 trols.

  1}  Linear Interpolation—In this
 method,  the (linear) calibration curve
 used to convert analyzer readings to
 concentration values is defined by a
 slope and intercept, which are up-
 dated at each calibration. Both unad-
 justed and adjusted response read-
 ings are required for each calibration.
 Each ambient concentration is calcu-
 lated from individual slope and inter-
 cept values determined by linear in-
 terpolation between the adjusted
 slope and intercept of the most re-
 cent previous calibration and the un-
 adjusted slope and intercept of the
 first subsequent calibration.
  Because of the need for subse-
 quent  (level 1) calibration informa-
 tion, this method cannot be used for
 real time calculation of concentration
 readings. Also, some contingency ar-
 rangement (such as method 2) must
 be employed when a subsequent cal-
 ibration is missing (e.g., following a
 disabling malfunction). Physical zero
 and span adjustments to the analyzer
 are needed only to maintain an ap-
 propriate scale range or to avoid
scale nonlinearity due to cumulative
drift in excess of design values.
  Within these constraints, data invali-
  dation limits should be based on net
  change from one calibration to the
  next, rather than on total cumulative
  drift, because the calibration is con-
  tinually updated.
    A significant problem with this
  method is acquiring the requisite cal-
  ibration data and  making sure it is
  merged correctly with the.ambient
  data to facilitate the required calcula-
  tions. Some automated data acquisi-
  tion systems support this application
  by making special provisions to ac-
  quire and process periodic zero and
  span data. One way to ensure that
  the zero/span data are correctly
  merged with the ambient readings is
  to code the zero and span values di-
  rectly into the data set at the location
  corresponding to the time of calibra-
  tion, replacing the normal hourly
  reading that is lost anyway because
  of the calibration.  This data can be
  marked (such as with a negative
  sign) to differentiate it from ambient
  data and later deleted from the final
  report printout.
    When  zero and span data is ac-
  quired automatically by.a  data acqui-
  sition system for direct computer
  processing, the system must be suffi-
  ciently sophisticated to:   ,
    a.  ensure that zero or span data
    is never inadvertently reported as
    ambient measurements;
    b.  ignore transient data during the
    stabilization period before the ana-
    lyzer has reached a stable zero or
    span response (this period may
    vary considerably from one ana-
    lyzer to another);
   c.   average the stable zero and span
    readings over some appropriate
   time period so that the zero or
   span reading obtained accurately
   represents the analyzer's true zero
   or span response;
   d.   ignore ambient readings for an
   appropriate period of time immedi-
   ately following a zero or span read-
   ing until the analyzer response has
   restabilized to the ambient-level
   concentration.

2)  Step-Change Update—This
method is similar to Method 1  above
except that the adjusted slope and
intercept of the most recent calibra-
tion are used to calculate all subse-
quent ambient readings until up-
dated by another calibration (i.e., no
interpolation). No unadjusted zero or
span readings are used, and ambient
measurements can be calculated in
real time if desired.  The same com-
ments concerning physical zero and
span adjustments and data  invalida-
  tion limits given for Method 1 apply,
  as well as the comments concerning
  zero and span data acquired auto-
  matically by a data acquisition sys-
  tem.

  3)  Major Calibration Update—In
  this method, the calibration slope
  and intercept used to calculate ambi-
  ent measurements are updated only
  for "major" calibration—i.e., monthly
  or quarterly multi-point calibrations.
  All ambient measurements are calcu-
  lated from the most recent major cal-
  ibration. Between major calibrations,
  periodic zero and span calibrations
  are used to measure the difference
  between the most recent major cali-
  bration and the current instrument
  response. Whenever this difference
  exceeds the established zero/span
  adjustment limits (see sections
  9.1.2 e and 9.1.2 f), physical zero and/
  or span adjustments are  made to the
  analyzer to restore  a match between
  the current analyzer response and
  the most .recent major calibration.
  Neither adjusted  nor unadjusted zero
  or span readings  are used in  the cal-
  culation of the ambient concentra-
  tions.

  4)  "Universal" Calibration—A fixed,
  "universal" calibration is established
  for the analyzer and used to calculate
  all ambient readings. All calibrations
  are used to measure the deviation of
 the current analyzer response from
 the universal calibration.  Whenever
 this deviation exceeds the estab-
 lished zero and span adjustment lim-
 its, physical zero and/or span  adjust-
 ments are made to the analyzer to
 match the current analyzer response
 to  the universal calibration.

 (b)  Invalidation of Ambient Data—
 When zero or span drift data valida-
•tion limits (see section 9.1.2 (f)) are '
 exceeded,  ambient measurements
 must be invalidated back to the most
 recent point in  time  where such mea-
 surements are known to be valid.
 Usually this point  is the previous cal-
 ibration (or accuracy audit), unless
 some other point in  time can be
 identified and related to the probable
 cause of the excessive drift (such as
 a power failure or  malfunction). Also,
 data following an analyzer malfunc-
 tion or period of non-operation
 should be regarded as invalid  until
the next subsequent (level 1) calibra-
tion unless unadjusted zero and span
readings at that calibration can
support its validity.
  Data quality assessment measure-
ments (precision and accuracy
checks) are not intended to be  used

-------
                            March 1985
                                                                         Section 2.0.9
 for data validation/invalidation. But if
 the assessment results clearly indi-
 cate a serious response problem with
 the analyzer, the agency should re-
 view all pertinent quality control in-
 formation to determine whether any
 ambient data, as well as any associ-
 ated assessment data, should be in-
 validated. However, this means
 should not be used merely to im-
 prove the data quality of specific an-
 alyzers or of the reporting agency.
   Procedures for screening data for
 possible errors or anomalies should
 also be implemented. Reference. 18
 recommends several statistical
 screening procedures for ambient  air
 quality data that should be applied to
 identify gross data  anomalies. Addi-
 tional information on validation of air
 monitoring data is contained in Ref-
 erences 19 and  20.

 (c)  Data Reporting—Procedures for
 coding, key punching, and data edit-
 ing should be documented and im-
 plemented. Recommended proce-
 dures for these  data processing
 activities are described in various
 sections of Reference 21: coding in-
 structions in Sections 3.4.1 through
 3.4.6; key punch instructions in Sec-
 tions 4.4.1 through 4.4.6; and data
 editing in Section 7.1.2.      •  ,

 (d)   Processing of  Data Quality As-
 sessment Information—It is of the ut-
 most importance that all precision
 and accuracy assessment readings
 from an analyzer be processed ex-
 actly as ambient readings recorded
 at that time would  be processed.
 Many automatic data acquisition and
 processing systems do not include •
 provision for handling such extra
 readings, and this capability  is diffi-
 cult to incorporate into such  systems
 unless it is done in the earliest plan-
 ning stage. External or hand process-
 ing of such readings should be dis-
 couraged unless it  is done with
 extreme care and assurance  that pro-
 cessing is identical to the way ambi-
• en.t readings are processed by the
 automatic system. Perhaps the best
 way to handle such readings is to  en-
 ter them into the automatic process-
 ing system in such a way that the
 system thinks they are actual ambi-
 ent readings and processes them ac-
 cordingly; After processing, the read-
 ings can be removed from the final
 ambient data listing and used in the
 data quality assessment calculations.

   When precision or accuracy as-
 sessment readings  are obtained dur-
 ing any period for which the ambient
 readings immediately before or im-  •
mediately after these readings are
determined by suitable reason to be
invalid, then the precision and accu-
racy readings should also be invali-
dated. Any data quality calculations
using the invalidated readings should
be redone. Also, the precision  or ac-
curacy checks should be resched-
uled, preferably in the same calendar
quarter. The basis or justification for
all data invalidations should be per-
manently documented.

9.1.4 Non-Programmed Adjust-
ments to Ambient Data—Adjust-
ments to ambient data as described
in the previous section, made rou-
tinely according to a documented,
pre-established procedure (pro-
grammed adjustments), would be a
normal part of an overall scheme to
maintain high levels of data quality.
In contrast, after-the-fact adjustments
or "corrections" are occasionally pro-
posed to ambient data based on
unanticipated events or discoveries.
This latter type of adjustment should
be scrutinized completely before any
changes are made to ambient  data.
In general, such adjustments are dis-
couraged as there is a substantial
risk that they may cause more harm
than good. There is also a risk that
such proposed adjustments might be
used or might appear to be used for
ulterior purposes. In many cases, this
type of correction may not be worth
the trouble of carrying it out.
  If, after scrutiny, a special, unpro-
grammed adjustment  is determined
to be appropriate and is made to a
block of ambient data, it is very im-
portant to ensure that the exact same
adjustment is also made to any pre-
cision and accuracy measurements
obtained during the affected time pe-
riod. Any data quality calculations af-
fected by the change should also be
recomputed. All such  adjustments
should be completely  documented,
including the rationale and justifica-
tion for the adjustment.

9.1.5 Written Operational Proce-
dures and Document Control—All
significant quality assurance proce-
dures should be described in writing'
in sufficient detail to assure that all
operators or analysts carry out the
procedures in the same way. Docu-
ment control should also be consid-
ered for these written  operational
procedures. Section 1.4.1 of Refer-
ence 4 provides information on es-
tablishing a document control  sys-
tem.
  As'outlined in Appendices A and 8
of 40 CFR Part 58, written  operational
 procedures should be available for at
 least the following monitoring activi-
 ties:
   1. selection of analyzers;
   2. training of analyzer operators;
   3. installation  of analyzers and as-
   sociated equipment;
   4. selection, Control, and traceabil-
   ity of calibration standards;
   5. procedures for multipoint  cali-
   brations;
   6. procedures for level 1 zero/span
   calibrations and adjustments of an-
   alyzers;
   7. procedures for .establishing the
   frequency of level 1 zero/span cali-
   brations (and level 2 checks, if
   used);
   8. control limits for zero, span and
   other control checks, and respec-
   tive corrective actions when such
   limits  are surpassed;
   9. calibration and zero/span checks
   for multiple range analyzers, if ap-
   plicable;
   10. preventive and  remedial
   maintenance;
  .11.  quality  control  procedures for
   air pollution episode monitoring;
   12.  data recording, processing, and
   validating procedures and limits;
   13.  data quality assessment (preci-
   sion and accuracy);
   14.  documentation  of quality  con-
   trol information.
 Guidance for many of these opera-
 tional procedures is currently avail-
 able in (a) analyzer operation or in-
 struction manuals, (b) EPA reference
 and equivalent methods, and (c)  EPA
 guideline documents, particularly
 References 7, 13, 14, 16, 17, 18, 21,
 and other  sections of this volume of
 the Handbook. However, it is the or-
 ganization's responsibility to develop
 its own  unique written operational
 procedures applicable to air quality
 measurements made by the organi-
 zation.

 9.1.6  Special Guidance for Epi-
 sodes—As defined here for the pur-
. pose of quality control, an air pollu-
 tion episode is any concentration
 equal to or greater than a  pollutant
 standard index (PSI) of 200. Pollutant
 concentrations corresponding to  PSI
 of 200 are shown in Tables 1 and 2
 of Reference 22.
   In addition to the previous guid-
 ance, the following procedures are
 recommended for analyzers used to
 monitor during air pollution
 episodes.
   1. A Level 1 zero and span calibra-
 tion should be performed  during the
 episode and at least weekly if the
 episode lasts longer than one week.

-------
                            Section 2.0,9
                                                        10
                                                                          March 1985
    2.  During  the episode season, but
  not during an episode, analyzers
  which are used to measure episode
  concentrations should be subjected
  to Level 1 zero and span calibrations
  at least every two weeks on the mea-
  surement range used for episode
  monitoring. This Level 1 zero and
  span calibration should be  performed
  at the same time the Level  1 zero
  and span calibration is performed on
  the analyzer's normal measurement
  range.
    3. During the episode season, but
  not during an episode, analyzers
  which have previously measured
  episode concentrations should be
  subjected to a performance audit of
  the type described in Appendices A
  and B.
    4. As soon as possible after an ac-
  tual episode the analyzer calibration
  should be checked with a 3 or more
  point performance audit, using
  standards different from the routine
  calibration standards.

  9.2   Quality  Assurance for
  Manual  Methods
   An appropriate and effective qual-
 ity control program for manual meth-
 ods is as necessary as it is for auto-
 mated analyzers, although the quality
 control activities will be more
 specific to the individual monitoring
 method  used. Suggestions and guid-
 ance for method-specific quality con-
 trol activities, such  as equipment per-
 formance and acceptance testing,
 calibration, and  data quality assess-
 ment procedures, are generally con-
 tained in the method description and
 in the method-specific sections of
 this Handbook (see sections 2.1, 2.2,
 2.4, and 2.8).

 9.2.1  Data Validation and Reporting
 (or Manual Methods—Monitoring
 data of poor quality may be  worse
 than no data at all.  For manual meth-
 ods, the first level of data validation
 should be to accept or reject moni-
 toring data based upon results from
 operational checks selected to moni-
 tor the critical  parameters in  all three
 major and distinct phases of manual
 methods—sampling, analysis, and
 data reduction.
  In addition to using operational
 checks for data validation, the user
 must observe all limitations,  accep-
 tance limits, and warnings described
 in the reference and equivalent meth-
ods per se that may invalidate data.
  It is further recommended  that re-
sults from performance audits re-
quired in  Appendices A and B not be
  used directly for data validation be-
  cause these checks (performance au-
  dits) are intended only to assess the
  quality of the data.
     Procedures for coding, key punch-
  ing, and data editing should be im-
  plemented. Recommended proce-
  dures for these data processing
  activities are described in Reference
  21: coding instructions in Sec-
  tions 3.4.1 through 3.4.6; key punch
  instructions in Sections 4.4.1 through
  4.4.6; and data editing in Sec-
  tion 7.1.2.
    Procedures for screening data  for
  possible errors or anomalies should
  also be implemented. References 18
  and 19 recommend several screening
  procedures for ambient air quality
  data that should be applied to iden-
  tify gross data anomalies.


  9.2.2  Written Operational Proce-
  dures and Document Control for
  Manual Methods—To standardize
  the approach used by different field
  operators and different laboratory
  analysts, certain operational  proce-
  dures (standard operating proce-
  dures, or SOP's) must be written  and
  readily available to all organization
  personnel. Document control  should
  also be considered for many of these
 written operational procedures. Sec-
 tion 1.4.1 of Reference 4 provides in-
 formation on  establishing a docu-
 ment control system. '
   Written operational procedures
 must be available for at least the fol-
 lowing TSP monitoring activities:
   1. calibration of the hi-vol sampler,
   2. important preventive mainte-
   nance tasks and a schedule  for
   completion of these tasks,
  3. calculation of monitoring con-
  centration, including an  example
  calculation,
  4. performance audits required in
  Appendices  A and B to assess ac-
  curacy,
  5. collocated sampling required in
  Appendices A and B to assess pre-
  cision,
  6. data  validation,
  7. field  handling of filters,
  8. flow measurements, and
  9. conditioning and  weighing of fil-
  ters.

  Written  operational procedures
must be available for at least the fol-
lowing SO2 and NO2 monitoring ac-
tivities:
  1. calibration of the air flow control
  devices,
  2. calibration of the spectrophoto-
  meter.
       •
    3. important preventive mainte-
    nance tasks and a schedule for
    completion of these tasks,
    4. calculation of monitoring con-
    centration, including an example
    calculation,
    5. performance  audits required in
    Appendix A to assess accuracy,
    6. collocated  sampling required in
    Appendix A to assess  precision,
    7. data validation,
    8. flow measurements, and
    9. chemical analysis..
 •   Written operational procedures
  must be available for at least the fol-
  lowing lead monitoring activities:.
    1. calibration  of the hi-vol sampler,
    2. calibration  of the atomic absorp-
    tion spectrophotometer,
    3. important preventive mainte-
    nance tasks and a schedule for
    completion of these tasks,
    4. calculation  of monitoring con-
    centration, including an example
    calculation,
    5. performance audits required in
    Appendices A and B to assess ac-
    curacy,
    6. collocated sampling required in
    Appendices A and B to assess pre-
    cision,
    7. data validation,
    8. flow measurement, and
    9. analysis.
    Many of these operational proce-
 dures are currently  in (a)  EPA refer-
 ence and equivalent methods, and
 (b) EPA guideline documents, partic-
 ularly this Handbook. However, it is
 the organization's responsibility to
 develop its own unique written oper-
 ational procedures applicable to air
 quality measurements made by the
 organization.
 9.3   References
 1.  Code of Federal Regulations, Title
 40, Chapter 1-, Part 58, "Ambient Air
 Quality Surveillance."
 2.  Code of Federal Regulations, Title
 40, Chapter 1, Part 58, Appendix A,
 "Quality Assurance Requirements for
 State and Local Air Monitoring Sta-
 tions (SLAMS)."
 3.  Code of Federal Regulations, Title
 40, Chapter 1, Part 58, Appendix B,
 "Quality Assurance Requirements for
 Prevention of Significant Deteriora-
 tion (PSD) Air Monitoring."
 4. "Quality Assurance Handbook for
 Air Pollution Measurement Systems,
 Volume l-Principles." EPA-600/9-75-'
 005. March 1976. Available from U.S.
 Environmental Protection Agency
ORD Publications 26 W. St. Clair St
Cincinnati, OH 45268.

-------
                           March 1988
                 11
Section 2.0.9
5. Code of Federal Regulations, Title
40, Chapter 1, Part .58, Appendix C,
"Ambient Air Quality Monitoring
Methodology."
6. Code of Federal Regulations, Title
40, Chapter 1, Part 53, "Ambient Air
Monitoring Reference and Equivalent
Methods."
7. U.S. Environmental Protection
Agency, "Ambient Monitoring Guide-
lines for Prevention of Significant De-
terioration (PSD)." EPA-450/2-78-019
(OAQPS 1.2-0.96). May 1978.

8. Kopecky, M.J. and B. Roger,
"Quality Assurance for Procurement
of Air Analyzers," 33rd Annual Tech-
nical Conference Transactions, Amer-
ican Society for Quality Control,
Houston, TX, May 1979.
9. Sexton, F.W., F.F. McEIroy, R.M.
Michie, Jr., V.L Thompson, and J.A.
Bowen.' Performance Test Results
and Comparative  Data1 for Designated
Reference and Equivalent Methods
for Ozone. EPA-600/4-83-003, U.S. En-
vironmental Protection Agency, Re-
search Triangle Park, NC 27711. April
1983.
10. Michie, P.M., Jr., F.F. McEIroy,
J.A. Sokash, V.L. Thompson, D.P.
Dayton, and,C.R. Sutcliffe. Perfor-
mance Test Results and Comparative
Data for Designated Reference Meth-
ods for Carbon Monoxide. EPA-600/
4-83-013, U.S. Environmental Protec-
tion Agency, Research Triangle Park,
NC 2771 I.June 1983.
11. Michie, R.M., Jr., F.F. McEIroy,
J.A. Sokash, V.L Thompson and B.P.
Fritschel. Performance Test Results
and Comparative Data for Designated
Reference and Equivalent Methods
for Nitrogen Dioxide. EPA-600/4-83-
019, U.S. Environmental Protection
Agency, Research Triangle Park, NC
27711. June 1983.
12. Michie, R.M., Jr., F.F..McEIroy,
F.W. Sexton, and V.L. Thompson.
Performance Test Results and Com-
parative Data for Designated Equiva-
lent Methods for Sulfur Dioxide. EPA-
600/4-84-015, U.S. Environmental
Protection Agency, Research Triangle
Park, NC 27711. January,  1984.
13. Ellis, E.G., "Technical Assistance
Document for the  Chemilumines-
cence Measurement of Nitrogen
Dioxide." EPA-600/4-75-003. U.S. En-
vironmental Protection Agency, Re-
search Triangle Park, NC 27711. De-
cember 1975.
14. Easton, W.C., "Use of the Flame
Photometric Detector Method for   •
Measurement of Sulfur Dioxide in
Ambient Air: A Technical Assistance
Document." EPA-600/4-78-024. U.S.
Environmental Protection Agency,
Research Triangle Park, NC 27711.
May 1978.
15. Von Lehmden, D.J., "Suppres-
sion Effect of C02 on FPD Total Sul-
fur Air Analyzers and Recommended
Corrective.Action." Proceedings, 4th
Joint Conference on Sensing Soci-
ety,  pp. 360-365, 1978.
16. Paur, RJ. and F.F. McEIroy.
Technical Assistance Document for
the Calibration of Ambient Ozone
Monitors. EPA-600/4-79-057. U.S. En-
vironmental Protection Agency, Re-
search Triangle Park, NC 27711. Sep-
tember 1979.

17. McEIroy, F.F. Transfer Standards
for the Calibration of Ambient Air
Monitoring Analyzers for Ozone.
EPA-600/4-79-056. U.S. Environmen-
tal Protection Agency, Research Tri-
angle Park, NC 27711. September
1979.
18. "Screening  Procedures for Ambi-
ent Air Quality Data." EPA-450/2-78-
037 (OAQPS 1.2-092). July 1978.
19. "Validation  of Air Monitoring
Data." EPA-600/4-80-030. U.S. Envi-
ronmental Protection Agency. June
1980.
20. Rhodes, R.C. "Guideline on the
Meaning and Use of Precision and
Accuracy Data Required by 40 CFR
Part  58, Appendices A and B." EPA-
600/4-83-023. U.S. Environmental
Protection Agency, Research Triangle
Park, NC 27711. June 1983.
21. "AEROS Manual  Series, Volume
II: AEROS Users Manual." EPA-450/
2-76-029 (OAQPS 1.2-0.39). December
1976.
22. Code of Federal Regulations,
Title 40, Chapter 1, Part 58, Appendix
G. "Uniform Air Quality Index and
Daily Reporting."

-------

-------
 Section 2.0.10   USEPA National Performance Audit Program
                             (February 1994)
 The National Performance Audit Program is a cooperative effort between EPA's
 Atmospheric Research and Exposure Assessment Laboratory (AREAL), the 10
 EPA  Regional Offices,  and the 170 state and local  agencies  that 'operate
 SLAMS/NAMS  air pollution  monitors.   Also included in the  NPAP are
 organizations that operate air monitors at PSD sites.  Participation in the NPAP
 is required for agencies operating SLAMS/NAMS and PSD monitors as p«r
 Section 2.4 of 40 CFR Part 58, Appendix A and Section 2.4 of 40 CFR Part 58
 Appendix B.  The NPAP is operated by the Quality Assurance Support Branch
 of AREAL.

 The goal of the NPAP is to provide audit materials and devices that  will enable
 o^AoaSSeSS the Proficiency of agencies who are operating monitors in the
 SLAMS/NAMS and  PSD networks.   To  accomplish this, the  NPAP  has
 established   acceptable  limits  or  performance   criteria,  based  on  the
 SLAMS/Nams and PSD requirements for each of the audit materials and devices
 provided  in  the program.   Any  device  or material  not meeting  the=e
'predetermined criteria is not used  in the program.    .

 All audit devices and materials used in the NPAP are certified as to their true
 value, and that certification is traceable to an NIST standard material or device
 wherever poss.ble. The audit materials used in the NPAP are as representative
 and comparable as possible to the calibration materials and actual air samples -
 used  and/or collected in the SLAMS/NAMS  and  PSD networks  The audit
 matenal/gas cylinder ranges used in the NPAP are those specified in the Federal
 register.

 The mailing address  and many of the audit  materials used in the National
 Performance Audit Program (NPAP) have changed since the publication of this
 section in 1979. The  present address is:

          Ambient Air Audit Coordinator
          Quality Assurance and Technical  Support Division
          Atmospheric Research and Exposure Assessment Laboratorv
          MD 77B
          Research Triangle Park, NC 27711

-------
The following audit materials are now used in the program:
Hi-Vo»/PM-10 (SSI)
The reference flow device (ReF) consists of a modified orifice, a wind deflector,
a manometer, and five resistance  plates.  The ReF for the PM-10 (SSI) flow
audit is similar except a filter is used as the only resistance.

Ozone
Ozone was added to the NPAP in 1989, The audit device is self-contained with
its own zero air and ozone generation system.

Dfchotomous (PM-10) (flow)
The dichot audit device consists of an inclined manometer filled with red gauge
oil, an altimeter that measures BP in millimeters, a small dial thermometer that
reads in °F, and the LFE (laminar flow element)  with air cleaner.  The dichot
measures fine flow (1 5.00 Ipm) and total flow (16.7 Ipm).

Lead (analysis)
The samples are 1.9 cm wide and 20 cm long glass fiber filter strips that have
been spiked with an aqueous solution of lead nitrate and oven-dried. Two filter
strips comprise a sample.

Sulfur Dioxide/NO-NO2/Carbon Monoxide (aas dilution system)
Beginning in 1991  one  gas  dilution system  was used  for all 3 audits. It.is
comprised of an audit device, one zero air system, and 2 cylinders of gas (NO2
and  a blend  of SO2/ NO, and CO).
Newly designed audit systems that have gas phase  titration capability are
presently being evaluated to determine their suitability for improving the stability
of NO2 audits. If acceptable, these devices will be phased in during the 1994
audit year.

Sulfate/Nitrate
The samples are 1.9 cm wide and 20 cm long glass fiber filter strips that have
been spiked with aqueous solutions of sodium sulfate and potassium nitrate and
oven-dried.  Three filter strips comprise a set.  This audit is voluntary since
sulfate/nitrate are not criteria pollutants.

-------
                           .  April 1885
                                                                        Section 2.0.11
                         11.0  Systems  Audit Criteria and Procedures
                             for Ambient Air  Monitoring Programs
 11.1   Introduction
 11.1.1  General - A systems audit is
 an on-site review and inspection of a
 state or local agency's ambient air
 monitoring program to assess its
 compliance with'established regula-
 tions governing the collection, analy-
 sis, validation, and reporting of ambi-
 ent air quality data. A systems audit
 of each state or autonomous agency
 within an EPA Region is performed.
 annually by a member of the Re-
 gional Quality Assurance (QA)  staff.
   The purpose of the guidance in-
 cluded here is to provide the regula-
 tory background and appropriate
 technical criteria which form the
 basis'for the air program evaluation
 by the Regional Audit Team. To pro-
 mote national uniformity in the eval-
 uation of state and local agency
 monitoring programs and agencies'
 performance, all EPA Regional  Of-
 fices are required to use at least the
 short form questionnaire (Section
 11.6), corrective action  implementa-
 tion request (CAIR) (Section 11.4.2),
 and the systems audit reporting for-
 mat (Section 11.4.4) each year. Use
 of sections of the long form ques-
 tionnaire is left to the discretion of
 the Regional QA Coordinator, with
 the concurrence  of the State or local
 agency. The short form questionnaire
 is essentially the same as the moni-
 toring  audit questionnaire used in
 FY-84. No substantive changes  have
 been made; however, the question-
 naire has been reorganized to im-
 prove the  information received  and
 facilitate its completion. In addition,
 requests for resubmission of data al-
 ready possessed by EPA have been
 deleted.
  The scope of a' system-s audit is of
 major concern to both EPA Regions
 and the agency to be evaluated. A
 systems audit as defined in the con-
 text of this document is seen to in-
 clude  an appraisal of the following
 program areas:  network manage-
 ment, field operations, laboratory op-
 erations, data management, quality
 assurance and reporting. The guid-
 ance provided concerning topics for
discussion during an on-site inter-
view have been  organized around
these key program areas (Section
 11.5). The depth of coverage within
these areas may be increased or de-
creased by using one or more sec-
tions of the long-form questionnaire
(Section 11.7) in conjunction with the
short-form questionnaire (Section
11.6). Besides the on-site interviews,
the evaluation should include the re-
view of some representative ambient
air monitoring sites and the monitor-
ing data processing procedure from
field acquisition through reporting
into the Storage And Retrieval Of Air
^ataJSAROAD) computer system.
  The systems audit results should
present a clear, complete and accu-
rate picture of the agency's acquisi-
tion of ambient air monitoring data.


11.1.2  Road Map to Using this Sec-
tion—This section contains guidance
and  sufficient information for operat-
ing a systems audit of an agency re-
sponsible for operating ambient air
monitoring sites, as part of the State
and  Local Air Monitoring Stations
(SLAMS) network, arid to report the
results in a uniform manner. The fol-
lowing topics are covered in the sub-
sections below:
  • a brief sketch of the regulatory
  requirements which dictate that
  systems audits be performed, indi-
  cating the regulatory uses to which
  the audit results may be  put (Sec-
  tion 11.2);
  • a discussion of
    1) the requirements on the
   agency operating the SLAMS net-
   work;
   2) program facets to be evalu-
   ated by the audit; and
   3) additional criteria to assist in
   determining the  required extent
   of the forthcoming audit (Section
   11.3);
  • a recommended audit protocol
  for use by the Regional Audit
  Team, followed by a detailed dis-
  cussion of audit results reporting
  (Section 11.4);
  • criteria for the evaluation of State
  and local agency performance in-
  cluding suggested topics  for dis-
  cussion during  the on-site inter-
  views (Section  11.5);
  • a short-form questionnaire,
  based  on the National Air Monitor-
  ing Audit Questionnaire prepared
  by  the STAPPA/ALAPCO Ad Hoc
  Air Monitoring Audit Committee.
  (10-20-83) (Section 11.6);
   • a long-form questionnaire, orga-
   nized around the six key program
   areas to be evaluated (Section
   11.7); and
   • a bibliography of APA guideline
   documents, which provides addi-
   tional technical background .for the
   different program areas under
   audit (Section 11.8).
 The guidance provided in this section
 is addressed primarily to EPA Re-
 gional QA Coordinators and mem-
 bers of the Regional audit teams to
 guide them in developing and imple-
 menting an effective and nationally
 uniform yearly audit program. How-
 ever, the criteria presented can also
 prove useful to agencies under audit
 to provide them with descriptions of
 the program areas to be evaluated.
   Clarification of certain sections,
 special agency circumstances,.and
 regulation or guideline changes may
 require additional discussion or infor-
 mation. For these reasons, a list of
 contact names and  telephone num-
 bers is given in Table 11-1.

 11.2  Regulatory Authority to
 Perform a Systems Audit

 11.2.1   General Regulatory Author-
 ity—The authority to perform sys-
 tems audits is derived from the Code
 of Federal Regulation (Title 40).
 Specifically: 40 CFR Part 35, which
 discusses agency grants and grant
 conditions, and 40 CFR Part 58,
 which deals specifically with the in-
 stallation, operation and quality as-
 surance of the SLAMS/NAMS net-
 works.
  The regulations contained in 40
 CFR Part 35 mandate the perfor-
 mance of yearly audits of agency air
 monitoring programs by the Re-
 gional Administrators or their de-
 signees. Pertinent regulatory cita-
 tions are summarized in Table 11-2.
 All citations are quoted directly from
 the regulations and  are intended as
 an indication of the  context within
 which systems audits are performed
 and the impact that  audit results may
 have on a given agency. Even
 though this is the regulatory author-
 ity to conduct such audits, for the
 SLAMS network, the specific author-
 ity is derived from 40 CFR Part 58.
Three specific citations from 40 CFR
 Part 58 are also, quoted in Table  11-2.

-------
                          Section 2.0.11
                                                    April 1986
 Ttblo 11-1.    List of Key Contacts and Telephone Numbers
  Assistance Area
  Office/Laboratory
      Name
  Telephone
   Number
EPA Location
 Laboratory        William J. Mitchell (919) 541-2769  EMSL/QAD/PEB
  Areas and NPAP                     FTS 629-2769

Monitoring
	 wnuam h, aarnora
WEIL FtfAWK.
Ctanlay 'C/om
{Witt 0*1 3UUb
(919) 541-5631
tlMiaUtatAUi'HtlU
O
OAQPS/MDAD/MRB
   Objectives/Siting
      Syetor,
                                     1Q1QLKA1
                                                    enact /A,i/\rt/riDO
 SAROAD
 System/NADP
Jake Summers
(919) 541-5694 OAQPS/MDAD/NADB
 NPAP = National Performance Audit Program
 PARS » Precision and Accuracy Reporting System
 NADB * National Aerometric Data Bank
  In addition to the regulations pre-
sented in Table 11-2, a further re-
quirement is imposed on reporting
organizations submitting data sum-
mary reports to the National Aero-
metric Data  Bank (NADB) through the
SAROAD computer system. SAROAD
acceptance criteria call for at least
75% data completeness, which has
been accepted as a data quality ob-
jective for state and local agencies'
monitoring  operations. The Regional
QA Coordinator may wish to  use this
requirement together with informa-
tion obtained by'accessing the
SAROAD NA285 or NA288 Computer
Programs, discussed in  Section 11.3.
The percent data completeness may
be effectively used as an indicator of
 Tabla 11-2.    Summary of Regulatory Authority to Conduct System Audits
    Section Number
    and Description
                                           Text
35.510-2

Grant Amount
35.510-3

Reduction in
Grant Amount

35.520

Criteria for (Grant)
Award
35.530
                       A.  Highlights of 40 CFR 35

   "In determining the amount of support for a control agency, the Regional Administrator will
   consider

   A.   The functions, duties and obligations assigned to the agency by an applicable
       implementation plan,

   B. the feasibility of the program in view of the resources to be made available to achieve or
      maintain EPA priorities and goafs,

   C. the probable or estimated total cost of the program in relation to its expected accomplish-
     ments,

   D. the extent of the actual or potential pollution problems,

   E. the population served within the agency's jurisdiction,

   F. the financial need, and,

   G. the evaluation of the agency's performance."

   "If the Regional Administrator's annual performance evaluation reveals that the grantee will
   fail or has failed to achieve the expected outputs described in his approved program, the
   grant amount shall be reduced..." •


   "No grant may be awarded to any interstate or intermunicipal air pollution control agency
   unless the applicant provides assurance satisfactory to the Regional Administrator that the
   agency provides for adequate representation of appropriate  State, interstate, local and (when
   appropriate) international interests in the air quality control region, and further that  the
   agency has the  capability of developing and implementing a comprehensive  air quality plan
   for the air quality control  region."

   "No grant may be awarded unless the Regional Administrator had determined that (1) the
   agency has the  capability or will develop the capability, to achieve the objectives and outputs
   described in its  EPA-approved program, and (2) the agency has considered and incorporated
   as appropriate the recommendations of the latest EPA performance evaluation in its pro-
   gram."

   In addition to any other requirements herein, each air pollution control grant shall be subject
   to the following conditions:

-------
                              April 1985
                                                                         Section Z.O.11
 Table 11-2.    Summary of Regulatory Authority to Conduct System Audits (Cont'd)
     Section Number
     and Description
                                          Text
  Grant Conditions
  35.538-1

  Agency
  Evaluation

  35.410

  Evaluation of
  Agency Perfor-
  mance
 A. Direct cost expenditures for the purchase of...

 B. The sum of non-Federal recurrent expenditures ...

 C. The grantee shall provide such information as the Regional Administrator may from time
    to time require to carry out his functions. Such information may contain, but is not limited
    to: Air quality data, emission inventory data, data describing progress toward compliance
    with regulations by specific sources, data on variances granted, quality assurance informa-
    tion related to data collection and analysis and similar regulatory motions, scfurce reduc-
    tion plans and procedures, real time air quality and control activities, other data related to
    air pollution emergency episode, and regulatory actions.

 "Agency evaluation ... should be continuous throughout the budget period. It is EPA policy to
 limit EPA evaluation to that which is necessary for responsible management of regional and
 national efforts to control air pollution. The Regional Administrator shall conduct an agency
 performance evaluation annually in accordance with 35.410."

 "A performance evaluation shall be conducted at least annually by the Regional Administrator
 and the grantee to provide a basis for measuring progress toward achievement of the ap-
 proved objectives and outputs described in the work program. The evaluation  shall be  consis-
 tent with the requirements of 35.538 for air pollution control agencies ..."
 Air Quality
 Surveillance
 Plan Content
 (SLAMS)
 58.23

 Monitoring
 Network
 Completion
 58.34

 NAMS Network
 Completion
 Appendix A
^Section 2.4

 National Perfor-
 mance and Sys-
 tems Audit
                     B.  Highlights of 40 CFR 58

 "By January 1, 1980 the State shall adopt and submit to the Administrator a revision to the
plan which will:                                                       .        .     .

A.  Provide for the ....

B.  Provide for meeting the requirements of Appendices A, C, D, and E, to this part

C.  Provide for the operation of...

D.  Provide for the review of the air quality surveillance system on an annual basis to deter-
    mine if the system meets the monitoring objectives defined in Appendix D to this part.
    Such review must..."

"By January 1, 1983:

A.  Each station in the SLAMS network must be in operation, be sited in accordance with  the
    criteria in Appendix E to this part, and be located as described on the station's SARD AD
    site identification form, and

B.  The quality assurance requirements of Appendix A to this part must be fully imple-
    mented."

"By January 1, 1981:

A.  Each NAMS must be in operation ...

B.  The quality assurance requirements of Appendix A to this part must be fully implemented
    for all NAMS."

"Agencies operating all or a portion of a SLAMS network are required to participate in EPA's
national performance audit program and to permit an annual EPA systems audit of their ambi-
ent air monitoring program ... for additional information about these programs.  Agencies
should contact either the appropriate EPA Regional Quality Control Coordinator  or the Quality
Assurance Branch, EMSL/'RTP, ... for instructions for participation."

-------
                         Section 2.0.11
                                  April 1985
whether a rigorous systems audit,
using the long form questionnaire,
might be needed or not.

11.2.2 Specific Regulatory Guidance
—The specific regulatory require-
ments of an EPA-acceptable quality
assurance program are to be found
in Appendix A to 40 CFR Part 58.
Section 2.2 of Appendix A details the
operations for which an agency must
have written procedures. The exact
format and organization of such pro-
cedures is not indicated, however.
Thus, many approaches to appropri-
ate documentation have been sug-
gested by EPA, local agencies and
other groups.
  One approach adopted by many
EPA Regional Offices is the organiza-
tion of the required material into the
framework recommended by the EPA
Quality Assurance Management Staff
in the document titled "Interim
Guidelines for the Preparation of
Quality Assurance Project Plans"
(QAMS 005/80, December 1980). The
sixteen (16) elements described in
the guideline document provide the
framework for organizing the re-
quired Air Program operational pro-
cedures, integrating quality assur-
ance activities and documenting
overall program operations. This ap-
proach is consistent with the re-
quired eleven items of 40 CFR
Part 58, Appendix A. Table 11-3 illus-
trates this consistency and demon-
strates how each required program
element will be evaluated in the con-
text of the program areas used in the
organization of the long-form ques-
tionnaire.

11.3   Guidelines for
Preliminary Assessment and
Systems Audit Planning
  In performing a systems audit of a
given  agency, the Regional QA Coor-
dinator is seeking a complete and ac-
curate picture of that agency's cur-
rent ambient air monitoring
operations. Past experience has
shown that four (4) person-days
should be allowed for an agency op-
erating 10-20 sites within close geo-
graphical proximity. The exact num-
ber of people and the time alloted to
conduct the audit are dependent on
the magnitude and complexity of the
agency and on the EPA Regional Of-
fice resources. During the alloted
time frame, the Regional QA Audit
Team should perform those inspec-
tions and interviews recommended
in Section 11.4. This includes on-site
interviews with key program person-
Table 11-3.    Specific Regulatory Requirements to be Evaluated in a
             Systems Audit
        Requirement
   (40 CFR 58, Appendix A)
 Pertinent Section
of OAMS Document
      005/80
  Pertinent Section
  of Questionnaire
       (11.7)
(1) Selection of Methods and
   Analyzers
(1) Selection of Methods,
   Analyzers
Project
Description
Organization &
Responsibility

QA Objectives

Sampling Proce-
dures
(11) Documentation of Quality   Sample Custody
    Control Information
(2) Installation of Equipment
(3) Calibration
(7) Calibration and Zero/Span
   Checks for Multiple Range
   Analyzers

   Only applicable if other
   than automated analyzers
   are used and analyses are
   being performed on filters •
   e.g., /VOJ or lead and TSP

(10) Recording and Validating
    Data
(4) Zero/span checks and ad-
   justments of automated
   analyzers
(5) Control checks and their
   frequency
(6) Control Limits for Zero/
   Span
(7) Calibration and Zero/Span
   for Multiple Range Ana-
   lyzers
(9) Quality control checks for
   air pollution episode
   monitoring

   Appendix A - Sections 2.0,
   3.0 and 4.0

(8) Preventive and Remedial
   Maintenance

   Appendix A - Section 4.0
(10)  Recording and Validating
     Data

(4) Zero/Span checks_and ad-
   justments of automated
   analyzers
Calibration
Procedures and
Frequency
Analytical
Procedures
Data Reduction,
Validation and
Reporting

Internal Quality
Control Checks
Performance and
System Audits

Preventive
Maintenance

Specific Routine
Procedures used
to Assess Data
Precision,
Accuracy and
Completeness
Corrective
Action
Planning
Planning
Planning
Planning

Field Operations


~Field/Lab Operations


Field/Lab Operations
Lab Operations





Data Management



Field/Lab Operations
                      QA/QC
 QA/QC
 Field/Lab Operations
 QA/QC
                      Data Management
 Field/Lab Operations

-------
                              April 1985
                                                                         'Section 2.0.11
  Table 11-3.    Specific Regulatory Requirements to be Evaluated in a
                Systems Audit
          Requirement
     (40 CFR 58, Appendix A)
  Pertinent Section
of OAMS Document
      005/80
  Pertinent Section
  of Questionnaire
       (11.7)
  (6) Control Limits and Cor-
     rective Actions

  (11) Documentation of Quality
      Control Information

  (W) Data Recording and
      Validation
Quality Assurance
Reports to
Management
Reporting
 nel, evaluations of some ambient air
 monitoring sites operated by the
 agency, and scrutiny of data process-
 ing procedures.

 11.3.1  Frequency of Audits—The
 EPA Regional Office retains the regu-
 latory responsibility to evaluate
 agency performance annually. Re-
 gional Offices are urged to use the
 short-form questionnaire (Section
 11.6), the CAIR (Fig. 11-4), and the
 audit reporting format (Section
 11.4.4 ). Utilizing the above to
 provide OAQPS with this audit infor-
 mation will establish a uniform basis
 for audit reporting throughout the
 country. For many well-established
 agencies, an extensive systems audit
 and rigorous inspection may not be
 necessary every year. The determina-
 tion of the extent of the systems
 audit and its rigor is left completely
 to EPA Regional Office discretion.
 Therefore, the option is provided
 here that extensive inspections and
 evaluations may be accomplished
 using the short-form questionnaire
 (Section 11.6), and appropriate sec-
 tion^) -of the long-form questionnaire
 (Section 11.7). It is suggested that a
 complete systems audit using the
 long-form questionnaire be per-
 formed at'least once every three
 years. Yearly reports must still, how-
 ever, include the short form, CAIR,
 and the report completed according
 to Section 11.4.4.
  The primary screening tools to  aid
 the EPA Regional QA Audit Team in
 determining which type of audit to
conduct and its required extent are:
   A. National Performance Audit
   Program (NPAP) Data—which  pro-
   vide detailed information on the
   ability of participants to certify
   transfer standards and/or calibrate
   monitoring  instrumentation. Audit
   Data summaries  provide a relative
   performance ranking for  each par-
  ticipating agency when compared
        to the other participants for a par-
        ticular pollutant. These data could
        be used as a preliminary assess-
        ment of laboratory operations at
        the different local  agencies.
      .  B. Precision and Accuracy Report-
        ing System (PARS) Data—which
        provide detailed information on
        precision and  accuracy checks for
        each local agency  and each pollu-
        tant, on a quarterly basis.  These
        data summaries could be  used to
        identify out-of-control conditions at
        different local  agencies, for certain
        pollutants.
        C. National Aerometric Data Bank
        (NADB) NA285 Data Summaries-7-
        which  provide a numerical count
        of monitors meeting and those not
        meeting specifications on  monitor-
        ing data completeness on  a quar-
        terly basis, together with an associ-
        ated summary of precision and
        accuracy probability limits. An ad-
        ditional program, NA288, will pro-
       vide data summaries indicating the
       percent of data by  site and/or by
       state for each pollutant.

     11.3.2  Selection of Monitoring Sites
     for Evaluation—It is suggested that
     approximately five percent (5%) of
     the sites of each  local agency in-
     cluded in the reporting organization
     be inspected during a systems audit.
     Many reporting organizations contain
     a large number of monitoring agen-
     cies,'while in other cases, a monitor-
     ing agency is its own  reporting orga-
     nization. For smaller local agencies,
     no fewer  than two (2) sites should be
     inspected. To insure that the  selected
     sites  represent a fair cross-section of
     agency operations, one half of the
     sites  to be evaluated should be se-
     lected by  the agency itself, while the
     other half should  be selected  by the
     Regional QA Audit Team.

       The audit team should use  both
     the Precision and Accuracy Reporting
     System (PARS) and the SAROAD
   computer databases in deciding on
   specific sites to be evaluated. High
   flexibility exists in the outputs
   obtainable from the NADB NA 285
   computer program; data
•   completeness can be assessed by
   pollutant, site,  agency, time period
   and season. These data summaries
   would assist the Regional audit team
   in spotting potentially persistent
   operational problems in need of
   more complete on-site evaluation. At
   least one site showing poor data
   completeness,  as defined by
   SAROAD, must be included in those
   selected to be evaluated.
    If the reporting organization under
   audit operates many sites and/or its
   structure is complicated and perhaps
   inhomogeneous, then an additional
   number of sites above the initial 5%
   level should be inspected so that a
  fair and accurate picture of the state
  and local agency's ability to conduct
  field monitoring activities can be ob-
  tained. At the completion of the site
  evaluations, the Regional audit team
 - is expected to have established  the
  adequacy of the operating proce-
  dures, the flow  of data from the sites
  and to be able to provide support to
  conclusions about the homogeneity
  of the reporting organization.

  11.3.3  Data Audits—With the imple-
  mentation by  many agencies of  auto-
  mated data  acquisition systems, the
  data management function has,  for
 the most  part, become increasingly
 complex.  Therefore, a  complete  sys-
 tems audit must include a review of
 the data processing and reporting
 procedures starting at  the acquisition
 stage and terminating  at the point of
 data entry into the SAROAD com-  •
 puter system. The process of audit-
 ing the data processing trail will  be
 dependent on  size and organizational
 characteristics of the repqrting orga-
 nization, the volume of data pro-
 cessed, and the data acquisition sys-
 tem's characteristics. The details  of
 performing a data processing audit
 are  left, therefore/to Regional and
 reporting organization personnel
 working together to establish a data
 processing audit trail appropriate for
 a given agency.
   Besides  establishing and docu-
 menting processing  trails, data pro-
 cessing audits  procedure must in-
 volve a certain amount of manual
 recomputation of raw data. The pre-
 liminary guidance provided here, for
 the number of data to be manually
 recalculated, should  be considered a
 minimum enabling only the detection
 of gross data mishandling:

-------
                          Section 2.0.11
                                                                         April 1985
   (a) For continuous monitoring of
   criteria pollutants, the Regional QA
   Coordinator should choose two 24-
   hour periods from the high and
   low seasons for that particular pol-
   lutant  per local agency per year. (In
   most cases the seasons of choice
   will be winter and summer). The
   pollutant and time interval choices
   are left to the  Regional auditor's
   discretion.
   (b) For manual monitoring, four
   24-hour periods per local agency
   per year should be recomputed.
   The Regional QA Coordinator
should choose the periods for the
data processing  audit while planning
the systems audit and ihspecting the
completeness records provided by
the NADB NA285 system. The recom-
mended  acceptance limits for the dif-
ferences between the data input into
SAROAD and that recalculated dur-
ing the on-site phase of the systems
audit, are given in Table 11-4.
   Systems audits conducted on large
reporting organizations  (e.g. four
local agencies) require recomputa-
tion of eight 24-hour periods for each
of the criteria pollutants monitored
continuously. This results  from  two
24-hour periods  being recomputed
for each  local agency, for each pollu-
tant monitored, during a given year.
For manual methods, sixteen 24-hour
periods are recomputed, consisting
of four periods per local agency, per
year.
11.4  Guidelines for
Conducting Systems Audits of
State and Local Agencies
  A systems audit should consist of
three separate phases :
  • Pre-Audit Activities
  • On-Site Audit Activities
  * Post-Audit Activities
  Summary activity flow diagrams
have been included as Figures 11-1,

Develop Audit Schedule

Contact Report
to Set Ter

Revise Schedu


ing Organizations
tative Dates

fe as Necessary


    Contact Reporting Organization to
       Discuss Audit Procedure
     Firm Dates for On-Site Visits
    Send Questionnaire and Request
     Preliminary Support Material
     Review Material Discuss with
   Reporting Organization QA Officer
      Develop Checklist of Points
            for Discussion
        Initiate Travel Plans
  Finalize Travel Plans with Information
   Provided by Reporting Organization
                       Contact Agency to Set Specific
                     Interview and Site Inspection Times
                                                                   Travel On-Site
Figure  11-1.   Pre-audit activities.

11-2 and 11-3, respectively. The
reader may find it useful to refer to
these diagrams while reading this
protocol.


11.4.1   Pre-Audit Activities - At the
beginning of each fiscal year, the Re-
Table 11-4.    Acceptance Criteria for Data Audits
Dafa Acquisition
Mode
Automatic Data
Retrieval
Stripchart
Records
Vlanuat
Reduction
Pollutants
SO2, QS, NOZ,
CO
SO2, O& NO2,
CO
TSP
Pb
Measurement
Range (ppm)(a>
0-0.5, or 0-1.0
0-20, or 0-50
0-0.5, or 0-1.0
0-20, or 0-50


Tolerance
Limits
±3 ppb
±0.3 ppm
±20 ppb
±1 ppm
±0.1 n.g/m3
WApptoptitta scaling should be used for higher measurement ranges.
tolSpecittad at 760 mm Hg and 25'C.
gional QA Coordinator or a desig-
nated member of the Regional QA
Audit Team, should establish a tenta-
tive schedule for on-site systems  au-
dits of the agencies within their re-
gion.
  Six (6) weeks prior to the audit, the
Regional QA Coordinator should con-
tact the Quality Assurance Officer
(QAO) of the reporting organization
to be audited to coordinate specific
dates and schedules for the on-site
audit visit. During this initial contact,
the Regional QA Coordinator should
arrange a tentative schedule for
meetings with key personnel  as well
as for inspection of selected ambient
air quality monitoring and measure-
ment operations. At the same time, a
schedule should be  set for the exit
interview used to debrief the  agency
Director or his designee, on the sys-
tems audit outcome. As part of this
scheduling, the Regional  QA Coordi-

-------
                              April 1985
                                    Section 2.0.11
             Audit Team Initial Interview of Reporting Organization Director
 'Audit Group 1
                         Interview with Key Personnel
                        Audit Group 2

Interview Planm

Interview Labors

Visit Labo
Witness Op

Review Sample R
Custot

Select Port/o
Initiate Auc


ng Manager

ory Director

ratory
erations

eceiving and
lY

n of Data
lit Trail

Establish Data Audit Trail
Through Laboratory Operations
to Data Management. Function




_ Meet to _
Discuss
Findings
i

Intervi
Operation

Visit Sites (Af,

Visit Sites (R

Visit Audit a
Fa

Select Por
Initiate A


w Field
s Manager

ency Selected)

eg/on Selected)

nd Calibration
ility

ion of Data
udit Trail

Establish Trail Through Field
Operations to Data Management
\
i
                    Finalize Audit Trails and Complete Data Audit
                         Prepare Audit Results nummary of
                   (a)  overall operations    tb)  data audit findings
                   lc)  laboratory operations  (d)  field operations
         Initiate Requests for Corrective Action Implementation Requests fCA/ft)
                  Discuss Findings with Key Personnel QA Officer
              Exit Interview with Reporting Organization Director to Obtain
             	Signatures on CAIR
                          On-Site Audit Complete
Figure  11 -2.   On-site activities.
nator should indicate any special re-
quirements such as access to specific
areas or activities. The Regional QA
Coordinator should  inform the
agency QAO that he will receive a
questionnaire, precision and accu-
racy data, and completeness "data
from NADB programs NA273 and
NA288 which is to be reviewed or
completed. He should emphasize that
the completed questionnaire is to be
returned to the EPA Region within
one (1) month of receipt. The addi-
tional information called for within
  e questionnaire is considered as a
  inimum, and both the Region and
  e agency under audit should feel
free to include additional informa-
tion.
  The Regional Audit Team may use
this initial contact or subsequent con-
versations to obtain appropriate
travel information, pertinent data on
monitoring sites to be visited, and
assistance in coordinating meeting
times.
  Once the completed questionnaire
has been received, it should be re-
viewed and compared with the  crite-
ria and information discussed in Sec-
tion 11.2 and with those documents
and regulations included by refer-
ence in Section 11.5.  The Regional
QA Audit Team should also use the
PARS and NADB NA273 and NA288
to augment the documentation  re-
ceived from  the reporting organiza-
tion under audit. This preliminary
 evaluation will be instrumental in se-
 lecting the sites to be evaluated and
 in the decision on the extent of the
 monitoring site data audit. The Re-
 gional Audit Team should then pre-
 pare a checklist detailing specific
 points for discussion with agency
 personnel.
   The Regional Audit Team could be
 made of several members to offer a
 wide variety of backgrounds and ex-
 pertise.  This team may then divide
 into groups once on-site, so that
 both audit coverage .and time utiliza-
 tion can be optimized. A possible di-
 vision may be that one group
 assesses the support laboratory and
 headquarters operations while
 another evaluates sites, and
 subsequently assesses audit and
 calibration information. The team
 leader should reconfirm the
 proposed audit schedule with the
 reporting organization immediately
 prior to travelling to the site.

 11.4.2  On-Site Activities—The Re-
 gional QA Audit Team should meet
 initially with the agency's Director or
 his designee to discuss the scope,
 duration, and activities involved with
 the audit. This should be followed by
 a  meeting with key personnel identi-
 fied from the completed question-
 naire, or indicated by the agency
 QAO. Key personnel to be inter-
 viewed_during the audit are those in-
 dividuals with responsibilities for:
 planning, field operations, laboratory
 operations,QA/QC, data manage-
 ment, and reporting. At the conclu-
 sion of these introductory meetings,
 the Regional Audit Team may begin
 work as two or more independent
 groups. A suggested auditing
 method is outlined  in Figure 11.2.
   To increase uniformity of site in-
 spections, it is suggested that a site
 checklist be developed and used.
   The importance of the data pro-
 cessing systems audit cannot be
 overstated. Thus, sufficient time and
 effort should be devoted to this activ-
 ity so that the audit team has a clear
 understanding and complete docu-
 mentation of data flow. Its impor-
 tance stems from the need to have
 documentation on the quality of am-
 bient air monitoring data for all the
 criteria pollutants for which the
 agency has monitoring requirements.
 The data processing systems audit
will serve as an effective framework
for organizing the extensive amount
of information gathered during the
audit of laboratory, field monitoring,
and support functions within the
agency.

-------
                         Saction 2.0.11
                                  April 1986
                    Travel Back to Regional Headquarters

                   	1	
I
1

1

1
1
1

1
1

r
I
1

!'
1
1

I
1
1

i

r
I
i
Audit Team Works Toget

Internal Review at Re

Incorporate Comments a

Issue Copies to Reporting
for Distribution and

her to Prepare Report \
-
gional Headquarters \
[

nd Revise Documents \
I

Organization Director
Written Comment

Incorporate Written Comments Received
from Reporting Organization
l
Submit Final Draft Report for
Internal Regional Review .
-~ ----- i - . -
1
I
Revise Report and Incorporate Comments
as Necessary
!
' 	 1
1
Prepare Final Copies
—-——-— 	 ,
1
Distribute to Reporting Organization
Director, OAQPS and Region
Flgun 11-3,   'Post-audit activities.

  The entire audit team should pre-
pare a brief written summary of find-
ings organized into the following
areas: planning, field operations, lab-
oratory operations, quality assur-
ance/quality control, data manage-
ment, and reporting. Problems with
specific areas should be  discussed
and an attempt made to  rank them in
order of their potential impact on
data quality. For the more serious of
these problems, Corrective Action
Implementation Request (CAIR)
forms should be initiated. An exam-
ple form is provided in Figure 11-4.

The forms have been designed such
that one is filled out for each major
deficiency noted that requires formal
corrective action.
  The format, content, and intended
use of CAIRs is fully discussed in
Section 11.4.5 of this document.
Briefly, they are request forms for
specific corrective actions. They are
initiated by the Regional QA Audit
Team and signed upon mutual
agreement by the agency's  Director
or his designee during the exit  inter-
view.
  The audit is now completed by
having the Regional Audit Team
members  meet once again with key
personnel, the QAO and finally with
the agency's Director or his designee
to present their findings. This is also
the opportunity for the agency to
present their disagreements. The
audit team should simply state the
audit results including an indication
of the potential data quality impact.
During these meetings the audit
team should also discuss the sys-
tems audit reporting schedule and
notify agency personnel that they
will be given a chance to  comment in
writing, within a certain time period,
on the prepared audit report in ad-
vance of any formal distribution.

11.4.3  Post-Audit Activities—The
major post-audit activity is the  prepa-
ration of the Systems Audit Report.
The report format is presented  in
Section 11.4.4.
  To prepare the report, the audit
team should meet and compare ob-
servations with collected  documents
and results of .interviews  and discus-
sions with key personnel. Expected
QA Project Plan implementation is
compared with observed  accomplish-
ments and deficiencies and the audit
findings are reviewed in detail.
Within thirty (30) calendar days of
the completion of the field work, the
audit report should  be prepared and
submitted.
  The Systems Audit Report is sub-
mitted to  the audited agency to-
gether with a letter thanking agency
personnel for their assistance, time
and cooperation. It is suggested that
the body of the letter be used to reit-
erate the fact that the audit report is
being provided for review and  writ-
ten comment. The letter should also
indicate that, should no written com-
ments be  received by the Regional
QA Coordinator within thirty (30) cal-
endar days from the report date, it
will be assumed acceptable to  the
agency in its current form, and will
be formally distributed without fur-
ther changes.
   If the agency has written  com-
ments or  questions concerning the
audit report, the Regional Audit
Team should review and  incorporate
them as appropriate, and subse-

-------
                                 April 1986
                                                                                 Section 2.0.11
                                         Corrective Action Implementation Request (CAIR)
  Reporting Organization.




  State or Local Agency	
  Deficiency Noted:
 Agreed-upon Corrective Action:
 Schedule for Corrective Action Implementation:
 Signed.
                                           Director.
                                           QA Officer.
                                           Audit Team Member.
                                        .Date




                                        .Date




                                        .Date
Corrective Action Implementation Report:
Signed.



Signed.
Director.
figure JJ-4.    Example of a CAIR Form.
                                          QA Officer.
                                        .Date



                                        .Date

-------
                         Section 2.0.11
                10
April 1985
quentty prepare and resubmit a re-
port In final form within thirty (30)
days of receipt of the written com-
ment. Copies of this report should be
sent to the agency Director or his de-
signe'e for his internal distribution.
The transmittal letter for the
amended report should indicate offi-
cial distribution and again draw at-
tention to the agreed-upon schedule
for Corrective Action Implementation.

11.4.4  Audit Reporting—The Sys-
tems Audit Report format discussed
in this section has been prepared to
be consistent with guidance offered
by the  STAPPA/ALAPCO Ad Hoc Air
Monitoring Audit Committee. The
format is considered as acceptable
for annual systems audit reports sub-
mitted  to the QAQPS. Regional Audit
Team members shall use this frame-
work as a starting point and include
additional material, comments, and
information provided by the agency
during  the audit to present an accu-
rate and complete picture of its oper-
ations and performance evaluation.
  At a minimum, the systems audit
report should include the following
six sections:
  Executive Summary—summarizes
the overall performance of the
agency's monjtoring program. It
should highlight problem areas need-
ing additional attention and should
describe any  significant conclusions
and/or  broad recommendations.
  Introduction—describes the pur--
pose and scope of the audit and_
identifies the Regional Audit Team
members, key agency personnel, and
other section or area leaders who
were interviewed. It should also
indicate the agency's facilities and
monitoring sites which were visited
and inspected, together with the
dates and times of the on-site audit
visit. Acknowledgement of the coop-
eration and assistance of the Director
and the QAO should also be consid-
ered for inclusion.
  Audit Results—presents sufficient
technical detail to allow a complete
understanding of the agency opera-
tions. The information obtained dur-
ing the audit  should be organized
using the recommended subjects and
the specific instructions given below.
It will be noted that the report format
follows the four-area organization of
the short-form questionnaire.

A. Network Design and Siting
   1} Network Size—Provide an
     overview of the network size
     and the number of local agen-
     cies responsible to the state for
     network operation.
  2)  Network Design and Siting—De-
     scribe any deficiencies in net-
     work design or probe siting dis-
     covered during the audit.
     Indicate what corrective actions
     are planned to deal with these
     deficiencies.
  3)  Network Audit—Briefly discuss
     the conclusions of the last net-
     work annual audit and outline
     any planned network  revision
     resulting from that audit.
  4)  Non-criteria Pollutants—Briefly
     discuss the agency's monitoring
     and quality assurance activities
     related to non-criteria pollu-
     tants.
B. Resources and Facilities
  1)  Instruments and Methods—De-
     scribe any instrument non-
     conformance with the require-
     ments of 40 CFR 50, 51, 53, and
     58. Briefly summarize agency
     needs for instrument  replace-
     ment over and above non-
   •  conforming instruments.
  2)  Staff and Facilities—Comment
     on staff training, adequacy of
     facilities  and availability of NBS-
     traceable standard materials
     and equipment necessary for
     the agency to properly conduct
     the bi-weekly precision checks
     and quarterly accuracy audits
     required under 40 CFR Part 58,
     Appendix A.
  3)  Laboratory Facilities—Discuss
     any deficiencies of laboratory
     procedures, staffing and facili-
     ties to conduct the tests and
     analyses needed to implement
     the SLAMS/NAMS monitoring
     the Quality Assurance, plans.
C. Data and Data Management
  1)  Data Processing and Submittal—
     Comment on the adequacy of
   .  the agency's staff and facilities
     to process and submit SAROAD
     air quality data as specified in
     40 CFR 58.35 and the  reporting
     requirements of 40 CFR  58, Ap-
     pendices A and F. Include an in-
     dication of the timeliness of
     data submission by indicating
     the fraction of data which are
     submitted more than  forty-five
     (45) days late.
  2)  Data Review—A brief discussion
     of the agency's performance in
     meeting  the 75% criteria for
     data completeness. Additionally,
     discuss any remedial  actions
     necessary to improve data re-
     porting.
  3)  Data Correction—Discuss the ad-
     equacy and documentation of
     corrections and/or deletions
     made to preliminary ambient air
      data, and their conistency with
      both the agency's QA Manual
      and Standard Operating Proce-
      dures, and any revised  proto-
      cols.
    4) Annual Report—Comment on
      the completeness, adequacy
      and timeliness of submission of
      the SLAMS Annual Report
      which  is required under 40 CFR
      58.26.
  D. Quality Assurance/Quality Control
    1) Status ofQuality Assurance Man-
      ual—Discuss the status of the
      Agency's Quality Assurance
      Plan. Include an indication of its
      approval status, the approval
      status  of recent changes and a
      general discussion of the con-
      sistency, determined during the
      systems audit, between the
      Agency Standard Operating Pro-
      cedures and the Quality Assur-
      ance Plan.
    2) Audit Participation—Indicate fre-
      quency of participation in an
      audit program. Include  as nec-
      essary, the agency's participa-
      tion in the  National Perform-
      ance Audit Program (NPAP) as
      required by 40 CFR Part 58.
      Comment on audit results and
      any corrective actions taken.
    3) Accuracy and Precision—As a
      goal, the 95% probability limits
      for precision (all pollutants) and
      TSP accuracy should be less"
      than ±15%. At 95% probability
      limits, the accuracy for all other
      pollutants should be less than
      ±20%. Using a short narrative
      and a summary table, compare
      the reporting organization's
      performance against these
      goals over the last two years.
      Explan any deviations.

    Discussion—includes a narrative of
  the way in which the audit results
  above are being interpreted. It
  should clearly identify the derivation
  of audit results which affect both
  data quality and overall agency oper-
  ations, and  should outline the basis
  in regulations and guideline docu-
  ments for the specific, mutually-
  agreed upon, corrective action rec-
  ommendations.
    Conclusions and Recommenda-
  tions—should center around the
  overall performance of the agency's
  monitoring  program. Major  problem
  areas should be highlighted. The
  salient facts of mutually agreed upon
  corrective action agreements should
  be included in this section. An
  equally important aspect to  be con-
  sidered in'the conclusion is  a deter-

-------
                              April 1985
                 11
                                   Section 2.0.11
  mination of the homogeneity of the
  agency's reporting organizations and
  the appropriateness of pooling the
  Precision and Accuracy data within
  the reporting organizations. The
  checklist in Figure 11 -5 should be
  included and submitted with the
  supporting documentation.
    Appendix of Supporting Documen-
  tation—contains a clean and legible
  copy of the completed short-form
  questionnaire and any Corrective Ac-
  tion Implementation Request Forms
  (CAIR). Additional  documentation
  may be included if it contributes sig-
  nificantly to a clearer understanding
  of audit results.

  11.4.5  Follow-up and Corrective Ac-
  tion Requirements—An effective cor-
  rective action procedure for use by
  the Regional  QA Audit Team follows.
  As a means of requesting corrective
  actions identified during the on-site
  audit, the auditor completes one
  copy of the form, shown in Figure
  11-4, for each major deficiency
  noted. These CAIR forms are pre-
  sented to, and discussed with, the
  agency's Director or his designee,
  and  its QAO during the-exit inter-
  view. Once agreement  has been
  reached, both the auditor and the Di-
  rector sign th'e form. The original is
  given to the agency Director or  his
  designee and a copy is retained by
  the auditor. A photocopy of the  com- •
  pleted CAIR is included in the audit
  report. It is taken to be  the responsi-
  bility of the agency to comply with
  agreed-upon corrective action re-
 quests in the  specified time frame.

 11.5  Criteria for the Evalua-
 tion  of State and Local
 Agency  Performance

  This section is designed to assist
 the Regional Audit Team in interpre-
 tation of the completed  questionnaire
 received back  from the agency prior
 to the on-site interviews. It also pro-
 vides the necessary guidance for top-
 ics  to be further developed during
 the on-site interviews.
  This section  is organized such that
 the specific topics to be  covered  and
 the appropriate technical guidance
 are keyed  to the major subject areas
 of the long-form questionnaire (Sec-
 tion 11.7). The left-hand  side of the
 page  itemizes the discussion topics
 (and the right-hand side provides  cita-
tions to specific regulations  and
guideline documents which  establish
the  technical background necessary
for the evaluation  of agency perform-
ance. A more complete bibliography
of EPA guideline documents is pre-
sented in 11.8.

-------
                          Section 2.0.11
12
April 1985
 11.5.1  Planning
                Topics for Discussion
                  Background Documents
  General information on reporting organization and
  status of Air Program, QA Plan and availability of
  SOPs
  Conformance of network design with regulation, and
  completeness of network documentation
  Organization staffing and adequacy of educational
  background and training of key personnel

  Adequacy of current facilities and proposed modifica-
  tions

 115.2  Field Operations
                Topics for Discussion	
  Routine operational practices for SLAMS network,
  and conformance with regulations
 • Types of analyzers and samplers used for SLAMS
  network
 * Adequacy of field procedures, standards used and
  field documentation employed for SLAMS network

 * Frequency of zero/span checks, calibrations and credi-
  bility of calibration equipment used

 • Traceability of monitoring and  calibration standards
 •  Preventive maintenance system including spare parts,
   tools and service contracts for major equipment
 *  Record keeping to include inspection of some site log
   books  and chain-of-custody procedures
 •  Data acquisition and handling system establishing a
   data audit trail from the site to the central data pro-
   cessing facility
   • State Implementation Plan
   • U.S. EPA QAMS 005/80
   • Previous Systems Audit report
   • QA Handbook for Air Pollution Measurement Sys-
     tems, Vol. II—Ambient Air Specific Methods, Section
     2.0.1.
   • 40 CFR 58 Appendices D and E
   • OAQPS Siting Documents (available by pollutant)
   • QA Handbook for Air Pollution Measurement Sys-
     tems, Vol. I—Principles, Section 1.4 Vol. II—Ambient
     Air Specific  Methods, Section 2.0.5
   	Background Documents	
   " QA Handbook for Air Pollution Measurement Sys-'
     terns. Vol. II, Section 2.0.9
   «.QA Handbook for Air Pollution Measurement Sys-
     tems, Vol. II
   « 40 CFR 50 plus Appendices A through G (potentially
     KforPM 10)
   « 40 CFR 58 Appendix C—Requirements for SCAMS an-
     alyzers
   • QA Handbook for Air Pollution Measurement Sys-
     tems, Vol. II
   « Instruction Manuals for Designated Analyzers
   • QA Handbook for Air Pollution Measurement Sys-
     tems, Vol. II—Ambient Air Specific Methods Section
     2.0.9
   • QA Handbook for Air Pollution Measurement Sys-
     tems, Vol. II—Ambient Air Specific Methods Section
     2.0.7
   » 40 CFR 58 Appendix A Section 2.3
   « QA Handbook for Air Pollution Measurement Sys-
     tems, Vol. II, Section 2.0.6
   « QA Handbook for Air Pollution Measurement Sys-
     tems, Vol. II—Ambient Air Specific Methods Sections
     2.0.3 and 2.0.9
11.5.3  Laboratory Operations
	Topics for Discussion	
• Routine operational practices for manual methods
  used in SLAMS network to include quality of chemi-
  cal and storage times
• List of analytical methods used for criteria pollutants
  and adherence to reference method protocols
• Additional analyses performed to satisfy regional,
  state or local requirements

• Laboratory quality control  including the regular usage
  of duplicates, blanks, spikes and  multi-point calibra-
  tions

• Participation in EPA NPAP and method for inclusion
  of audit materials in analytical run
  Documentation and traceability of laboratory mea-
  surements such as weighing, humidity and tempera-
  ture determinations
                 Background Documents
    40 CFR 50 Appendices A and B, and QA Handook,
    Vol. II

    40 CFR 58 Appendix C; "List of Designated Reference
    and  Equivalent Methods"
    Refer to locally available protocols for analysis of
    aldehydes, sulfate, nitrate, pollens, hydrocarbons,  or
    other toxic air contaminants.
    U.S. EPA APTD-1132 "Quality  Control Practices in
    Processing Air Pollution Samples"
    40 CFR 58 Appendix C; "List of Designated Reference
    and  Equivalent Methods"
    40 CFR 58 Appendix A Section 2.4
    QA Handbook for Air Pollution Measurement Sys-
    tems, Vol. II, Section 2.0.10
    40 CFR 58 Appendix C; "List of Designated Reference
    and  Equivalent Methods"
    40 CFR 58 Appendix C; "List of Designated Reference
    and  Equivalent Methods"

-------
                             April 1985
                                                      13
                                                                        Section 2.O.11
 •  Preventive maintenance in the laboratory to include
   service contracts on major pieces of instrumentation
 •  Laboratory record keeping and chain-of-custody pro-
   cedures to include inspection of logbooks used
 •  Adequacy of Laboratory facilities. Health and Safety
   practices and disposal of wastes

 •  Data acquisition, handling and manipulation system
   establishing data flow in the laboratory, data back-up
   system and data reduction steps
 •  Data validation procedures, establishing an audit trail
   for the laboratory to the central data processing facil-
   ity

 11.5.4  Data Management
 	         Topics for Discussion	
 •  Data flow from field and laboratory activities to cen-
   tral data processing facility
 •  Extent of computerization of data management sys-
   tem and verification of media changes, transcriptions
   and manual data entry
 •  Software used for processing and its documentation;
   to include functional description of software, test
   cases and configuration control for subsequent revi-
   sions
 •  System back-up and recovery capabilities
 •  Data screening, flagging and validation
• Data correction procedures and key personnel al-
  lowed to correct ambient air data
• Reports generated for in-house distribution and for
  submittal to EPA
• Responsibiffty for preparing data for entry into the
  SAROAD and PARS systems and for responsibility for.
  its final validation prior to submission

11.5.5 QA/QC Program
   	      Topics for'Discussion	
• Status of QA Program and its implementation
• Documentation of audit procedures, integrity of audit
  devices and acceptance criteria for audit results
• Participation in the National Performance Audit Pro-
  gram for what pollutants and ranking of results
• Additional internal audits such as document reviews
  or data.processing audits
• Procedure and implementation of corrective action
• Frequency of performance and concentration levels
  for precision checks for each criteria pollutant

11.5.6  Reporting
	   Topics for Discussion	
• Preparation of precision and accurancy summaries
  for the  PARS system
• Other internal reports used to track performance and
  corrective action implemenation
• Summary air data reports required by regulations
• Completeness, legibility and validity of P & A data on
  Form 1
 • 40 CFR 58 Appendix C; "List of Designated Reference
  and Equivalent Methods"
 • QA Handbook for Air Pollution Measurement Sys-
  tems, VoL II, Section 2.0.6
 • Handbook for Analytical Quality Control in Water and
  Wastewater Laboratories

  QA Handbook for Air Pollution Measurement Sys-
  tems, Vol. II, Sections 2.0.3 and 2.0.9

  Annual Book of ASTM Standards, Part 41, 1978.
  Standard Recommended Practice for Dealing with
  Outlying Observations (E 178-75)


               Background Documents
  QA Handbook for Air Pollution Measurement Sys-
  terns. Vol. II, Section 2.0.3
  QA Handbook for Air Pollution Measurement Sys-
  tems, Vol. II, Section 2.0.9

  QA Handbook for Air Pollution Measurement Sys-
  tems, Vol. II, Sections 2.0.3 and 2.0.9
  Validation of Air Monitoring Data, EPA-600/4-80-030
  Screening Procedures for Ambient Air Quality Data,
  EPA-450/2-78-037
  QA Handbook for Air Pollution Measurement
  Systems, Vol. II, Section 2.0.9
•. Aeros Manual Series, Vol. II, Aeros User's Manual,
  EPA-450/2-76-029                         .
     	Background Documents	
  40 CFR 58 Appendix A and QAMS 005/80
  QA Handbook for Air Pollution Measurement Sys-
  tems, Vol. II, Sections 2.0.11 and 2.0.12
  40 CFR 58 Appendix A
  QA Handbook for Air Pollution Measurement Sys-
  tems, Vol. II, Section 2.0.10
• 40 CFR 58 Appendix A
  	         Background Documents
• PARS User's Manual (in preparation)
• 40 CFR 58 Appendix A
  40 CFR 58 Appendices F and G
  40 CFR 58 Appendix A

-------
                             Section 2.0.11                   14                  April 1985
                                          Reporting Organization Homogeneity Checklist








                                                                                                              Yes      No



 1,  field operations, for all local agencies, conducted by a common team of field operators?                          	    	




 2.  Common calibration facilities are used for all local agencies?                                                  	    	



 3,  Precision checks performed by common staff for all local agencies?                                             	    	




 4.  Accuracy checks performed by common staff for all local agencies?                                             	    	



 5.  Otta handling follows uniform procedures for all local agencies?                                               j	    	



 6.  Central data processing facilities used for all reporting?                               '                         _;	    	




 7.  Tracaabllity of all standards established by one central support laboratory?                                      	  '  _,	




 5.  One central analytical laboratory handles all analyses for manual methods?        •                             	    	
Figure 11-5.    Example of Reporting Organization Homogeneity Checklist

-------
                              April 1986
                                 15
            Section 2.0.11
 11.6   Systems Audit
 Questionnaire (Short-Form)
  The short-form questionnaire has
 been designed specifically for use in
 annually reviewing state and local
 agencies' air monitoring programs. If
 the Regional QA Coordinator decides
 that a more rigorous systems audit
 and site inspections are necessary,
 he can utilize appropriate section(s)
 of the Long-Form Questionnaire
 (Section 11.7). This questionnaire has
                been designed around the format
                recommended by STAPPA/ALAPCO
                in the National Ambient Air Monitor-
                ing Questionnaire and is organized
                around four (4) major topics consis-
                tent with the reporting format out-
                lined in Section 11.4.4. They are:
                  A. Network Design and Siting
                  B. Resources and Facilities
                  C. Data Management, and
                  D. Quality Assurance and Quality
                  Control
                                    NATIONAL AIR MONITORING SYSTEMS AUDIT
                                                 QUESTIONNAIRE

                                                 (SHORT FORM)
 Agency,
 Address.
 Telephone Number (Area Code).
                 Number.
 Reporting Period (beginning-ending dates).

 Organization Director	
Air Program Supervisor.
 Data Management Supervisor.

 Quality Assurance Officer	

 Questionnaire Completed	
 On-Site Visit
 Date:	
         (date)


. Audit Team Members:
                                                                            (by)
Affiliation of Audit Team .
                                          SHORT FORM QUESTIONNAIRE
                                              TABLE OF CONTENTS


                                           Page No.
A.  NETWORK DESIGN AND SITING
   1. Network Size
   2. Network Design and Siting
   3. Network Review
   4. Non-Criteria Pollutants

B.  RESOURCES AND FACILITIES
   1. Instruments and Methods
   2. Staff and Facilities
   3. Laboratory Operations and Facilities
   4. Standards and Traceability

C.  DATA AND DATA MANAGEMENT
   1. Timeliness of Data
   2, Data Review
   3. Data Correction
   4. Annual Report
                      SF-2
                      SF-3
                      SF-3
                      SF-4
                      SF-4
                      SF-5
                      SF-5
                      SF-6
                      SF-7
                      SF-8
                      SF-8
                      SF-3
D. QUALITY ASSURANCE/QUALITY CONTROL
  1. Status of Quality Assurance Program
  2. Audit Participation
  3. Precision and Accuracy Goals •
SF-9
SF-9
SF-rO
                                                    SF-1

-------
                         Section 2.0.11                18                April 1985

                                         A. NETWORK DESIGN SITING
 7. NETWORK SIZE
 (a)  Complete the table below for each of the criteria pollutants monitored as part of your air monitoring network,
     Include only those sites that are presently operating and those which are temporarily inoperative (off line less
     than 30 days}. Do not include additional monitors which are collocated or index sites.
                                              Number of Monitors
                           SO2         NO2         CO          O3          fSP         Pb
     NAMS
     SLAMS
     (excluding NAMS)
    SPM
    TOTAL
(b)  SLAMS Network Description
     1. Whatls tha date of the most current official SLAMS Network Description?	;	
     2. Whara is it available for public inspection?:	   '	
     3. Does It include for each site the following?
                                                                                       YES         NO
     SAROAD Site /D#	
     Location                                   .                                      	     	
     Sampling and Analysis Method                                                           .	
     Operative Schedule	
     Monitoring Objective and Scale of Representativeness                                 	     	
     Any Proposed Changes                                                            	     	
(c)  For each of the criteria pollutants, how many modifications (SLAMS including NAMS) have been made since the
    last systems audit? (List the total SLAMS and NAMS).
    Date of last systems audit.
                                                 SF-2

-------
                           April 1985
                                                  17
                                                                    Section 2.0.11
Pollutant
Sulfur Dioxide
Nitrogen Dioxide
Carbon Monoxide
Ozone
                                      Added
                                                          Number of Monitors
                                             Deleted
                                                                                     Relocated
    Lead
(d)  Briefly discuss changes to the Air Monitoring Network planned for the next audit period. (Equipment is discussed
    in Part B).                                .              -
2. NETWORK DESIGN AND SITING
Indicate by SAROAD Number any non-conformance with the requirements of 40 CFR 58, Appendices D and E.
Monitor
Site ID (SAROAD)
                                                       Reason for Non-Conformance
 SO,
 CO
 NO2
 f&f
 PM-IO
 Pb
3.  NETWORK REVIEW
Please provide the following information on your previous internal Network Review required by 40 CFR 58.20d.
Review performed on:  Date: _ __ _ _______ _
Performed by: _ _ _ __ _ _
Location and Title of Review Document:
Briefly discuss all problems uncovered by this review.
                                                SF-3

-------
                         8«otlon 2.0.11                18               April 1985


 4, NON-CRITERIA POLLUTANTS

 Does your agency monitor and/or analyze for non-criteria and/or toxic air pollutants?  Yes	  No	

 If ysst please complete the form below.



         •                              .     Monitoring                                    SOP Available
 Pollutant                                 Method/Instrument                                     Yes/No
                                       B. RESOURCES AND FACILITIES

 1. INSTRUMENTS AND METHODS

 (a)  Please complete the table below to indicate which analyzers do not conform with the requirements of 40 CFR 53
     for NAMS, SLAMS, or SIP related SPM's.



     _ „                                                            Site               Comment on
     Pollutant	Number           Make/Model           Identification            Variances


      CO



     -SO2  "                                           '.'••'.



     NO2



     03


     pH-J»
     Pb
(b)  Please comment briefly on your currently identified equipment needs.
                                                 SF-4

-------
                            ApriHSSS                19               Section 2.0.11

 2. STAFF AND FACILITIES
 (a)  Please indicate the number of people available to each of the following program areas:

                                                                        Comment on Need for
     Program Area                              Number                  Additional Personnel
     Network Design and Siting

     Resources and Facilities

     Data and Data Management
     QA/QC
 (b) Comment on your agency's need for additional physical space (laboratory, office, storage, etc.).
3. LABORATORY OPERATION AND FACILITIES
(a)  Is the documentation of Laboratory Standard Operating Procedures complete?  Yes   •     No
    Please complete the table below.

    Analysis                           .                    Date of Last Revision
     Sfc
10
     Pb

     S04

     NO3

     SO2
            (bubblers)
     NO2

     Others (list by pollutant)
(b)  Is sufficient instrumentation available to conduct your laboratory analyses?  Yes        No
                                                  SF-5

-------
                         Section 2.0.11
                          20
                                                                      ApriM985
 If no, please Indicate instrumentation needs in the table below.
 Instrument
 Needed
Analysis
  New or
Replacement
  Year of
Acquisition
 4. STANDARDS AND TRACEABILITY

 (a) Please complete the table for your agency's laboratory standards.
 Parameter
      Primary
      Standard
    Secondary
     Standard
 CO
N02
SO2
Weights
Temperature
Moisture
Barometric Pressure
Flow
Lead
Other (specifiy)
Recertification
    Date
                                                  SF-6

-------
                             April 1985
                              21
                                  Section 2.0.11
(b)  Please complete the table below for your agency's site standards (up to 7% of the sites, not to exceed 20 sites).
    Parameter
              Primary
              Standard
                      Secondary
                       Standard
                                                                                      Recertification
                                                                                          Date
    CO .
    NO,
    SO2
                                       C. DATA AND DATA MANAGEMENT
T.  TIMELINESS OF DATA
For the current calendar year or portion thereof which ended at least 135 calendar days prior to the receipt of this
questionnaire, please provide the following percentages for required data submitted.

                                            % Submitted on Time*
Monitoring
   Qtr.
SO2
CO
NO,
                                                         PH-lo
                                                                                                Pb
1 (Jan. 1-March 31)
2 (Apr. 1-June 30)
3 (July 1-Sept. 30)
4 (Oct. 1-Dec. 31)
•"On-time" = within 135 calendar days after the end of the quarter in which the.data were collected.
                                                    SF-7

-------
Section 2.0.1 1
                                                      22
                                                                        April 1 985
  2. DATA REVIEW
  What fraction of the SLAMS sites (by pollutant) reported less than 75% of the data (adjusted for seasonal monitorinq and
  site start-ups and terminations)?
  Calendar Year

Pollutant
Percent of Sites
< 75% Data Recovery
/si 2nd 3rd 4th
Quarter Quarter Quarter Quarter
Ozone
Nitrogen Dioxide
Sulfur Dioxide
Carbon Monoxide



Lead .. - . .
 3. DATA CORRECTION
 (a)  Are changes to submitted data documented in a permanent file?
     Yes 	   No  	
     // no, why not?	  	
anlePnjeft
aCC°rding tO a
                                                   Standard Operating Procedure or your Agency Quality Assur-
    Yes
                        NO
    If not according to the OA Project Plan, please attach a copy of your current Standard Operating Procedure.
(c)  Who has signature authority for approving corrections?

                    (name}                                         (Program Function)           "
4. ANNUAL REPORT
(a)  Please provide the dates annual reports have been submitted in the last two years.
                                                  SF-8

-------
                            April 1986                23                Section 2.0.11

(b)  Does the agency's annual report fas required in 40 CFR 58.26) include the following?

                                                                                       YES          NO
    1.  Data summary required in Appendix F.
    2.  Annual precision and accuracy information described in Section && of Appendix A.
    3.  Location, date, pollution source and duration of all episodes reaching the significant
       harm levels.
    4.  Certification by a senior officer in the State or his designee.
(c)  Describe any deficiencies which cause the answer to part (b) of this question to be No.
                                  D. QUALITY ASSURANCE/QUALITY CONTROL
1.  STATUS OF QUALITY ASSURANCE PROGRAM
(a)  Does the agency have an EPA-approved quality assurance program plan?*
    Yes	  No	
  •' If yes, have changes to the plan been approved by the EPA?
    Yes	  No	
    Please provide:
    Date of Original Approval	
    Date of Last Revision	
    Date of Latest Approval.
(b)  Do you have any revisions to your QA Program Plan still pending?
    Yes        No
2.  AUDIT PARTICIPATION
(a)  Date last systems audit was conducted?.
    By whom?	'
(b)  Does the agency participate in the National Performance Audit Program (NPAP) as required under 40 CFR 58
    Appendix A?*
 •   Yes         No
*lf answer is No, give a brief summary of the deficiencies.
                                                  SF-9

-------
                          Saetlon 2.0.1 1                24                April 1985


 (c)  Please complete the table below.


     Parameter Audited                                          Date of Last NPAP Audit


     SO2 (Continuous)
     CO
     Pb
     fJeF Qovioo
     GOjfbubbhrl

     /V 0A ( COAT/AJ u o us-)

     ft/Q. fhsihh.'A. 1
     I " ^•'JC 1 UQtJiJt Ol /
3. PRECISION AND ACCURACY GOALS'
  As a goal, the 95 percent probability limits for precision (all pollutants) and TSP accuracy should be less than + 75
psrcenf. 
-------
                             April 1985                25                 Section 2,0.11


(b) Accuracy Goals

                                #  of Reporting                             Precision
    Pollutant                    Organization                     Qtr/Yr  Qtr/Yr  Qtr/Yr  Qtr/Yr
    NO2
    SO,
    CO
    f&P-
    Pb
(c) To the extent possible, describe problems preventing the meeting of precision and accuracy goals.
                                                   SF-11

-------
                           Section 2.0.11
                                                      26
                                                                        April 1986
 11.7  Systems Audit Ques-
 tionnaire (Long-Form)


   The long-form systems audit ques-
 tionnaire which follows is intended
 to provide a complete picture of
 agency ambient air monitoring oper-
 ations and quality assurance imple-
 mentation. J[h_e_followin£ Instructions
 might prove helpTfurin completing
 this survey questionnaire.
    1. For ease in completing the ques-
      tionnaire, it is not necessary to
      type. Filling it out legibly in black
      Ink is acceptable.
    2. Feel free to elaborate on any point
      or question in the form. Use addi-
      tional pages as necessary to give
      a complete response.

    3. When necessary, include copies of
      documents which will aid in un-
      derstanding your response.
4.  Please pay careful attention in
   cornjpleting the questionnaire. The
   information supplie'd will have~a
   direct bearing on the conclusions
   drawn and recommendations
   made concerning the evaluation
   of your organization's program.
5.  The Regional Quality Assurance
   Coordinator or a member, of his
   staff may be contacted for assis-
   tance in completing the question-
   naire.
                                  SYSTEMS AUDIT QUESTIONNAIRE (LONG FORM)

                                            GENERAL INFORMATION
 Questionnaire completion date.

 On-site systems audit date	

 Reporting period	
 Agency name and address.
 Mailing address (if different from above).
 Telephone number (FTS).

 Commercial (	)	

 Agency Director	
 Agency QA Officer	

 Reporting organizations making up this agency.
 Systems audit conducted by.

 Affiliation of audit team	
 Key Personnel:

'Planning	
Completed Questionnaire
   Interviewed
Field Operations.
 Laboratory Operations.

 QA/QC	
Data Management.

Reporting	
Persons Present during exit interview
                                                   LF-1

-------
                            April 1985                27                Section 2.0.11
            LONG FORM QUESTIONNAIRE

               TABLE OF CONTENTS

                                       PAGE NO.

A. NETWORK MANAGEMENT

   1.  General                            LF-3
   2.  Network Design and Siting           LF-5
   3.  Organization, Staffing and Training    LF-7
   4.  Facilities                            LF-8

B. FIELD OPERATIONS

   1.  Routine Operations                  LF-9
   2.  Quality Control                  '    LF-11
   3.  Preventive Maintenance              LF-14
   4.  Record Keeping                     LF-15
   5.  Data Acquisition and Handling        LF-16

C. LABORATORY OPERATIONS

   1.  Routine Operations                  LF-17
   2.  Quality Control                      LF-19
   3.  Preventive Maintenance              LF-21
   4.  Record Keeping                     LF-22
   5.  Data Acquisition and Handling        LF-23
   6.  Specific Pollutants
        TSP                              LF-23
        Lead     '                        LF-25

D. DATA ANB DATA MANAGEMENT

   7.  Data Handling                       LF-26
   2.  Software Documentation              LF-27
   3.  Data Validation and Correction        LF-28  •
   4.  Data Processing                     LF-29
   5.  Internal Reporting                    LF-32
   6.  External Reporting                   LF-33


E.  QUALITY ASSURANCE/QUALITY CONTROL
        «
   7.  Status of Quality Assurance Program   LF-35
   2.  Audits and Audit System  Trace-
      ability                              L.F-3S
   3.  National Performance Audit Pro-
      gram (NPAP) and Additional
      Audits                              LF-37
   4.  Documentation and Data Process-
      ing Review                          LF-38
   5.  Corrective Action System              LF-39
   6. Audit Result Acceptance Criteria       LF-39
                                                  LF-2

-------
                          Section 2.0.11                28                April 1985

                                          A. NETWORK MANAGEMENT
 7. GENERAL
 (a)  Provide an organization chart clearly showing the agency's structure and its reporting organizations. (Attach
     sheet(s) as necessary.)
 (b)  What is the basis for the current structure of the agency's reporting organizations?
                                                                                              Yes      No
     Field operations for all local agencies, conducted by a common team of field operators?        	    •	
     Common calibration facilities are used for all local agencies?                                	    	
     Precision checks performed by common staff for all local agencies?                          	    	
     Accuracy checks performed by common staff for all local agencies?                          	•_    	
     Data handling follows uniform procedures for all local agencies?
     Central data processing facilities used for all reporting?
     Traceability of all standards established by one central support laboratory?                  .  	    	
     One central analytical laboratory handles all analyses for manual methods?                    _^___    	
 (c)  Does the agency feel that the data for the reporting organizations it contains can be pooled?
     Yes        No         Please comment on either answer	
(d) Briefly describe any changes which will be made within the agency's monitoring program the next calendar year.
(e)  Complete the table below for each of the criteria pollutants monitored as part of your air monitoring network.
                                                          Number of Monitors
    •	:	P*-'0	
                                   SOZ         NO2          CO           O3          ?&>          Pb

    NAMS
    SLAMS (excluding NAMS)
    SPM
    TOTAL
(f)  What is the date of the most current official SLAMS Network Description?.
    /.  Where is it available for public inspection?.
                                                    LF-3

-------
                            April 1985                29                Section 2.0.11

 //. Does it include for each site the following?
                                                                  YES     NO
 SAROAD Site /D#  .                                               	   	
 Location                                                         	   	
 Sampling and Analysis Method                                     	   	
 Operative Schedule                                               	   	
 Monitoring Objective and Scale of Representativeness                	    	
 Any Proposed Changes                                            	   	
 (g)  For each of the criteria pollutants, how many modifications (SLAMS including NAMS) have been made since the
     last systems audit? (List the total SLAMS and NAMS)
     Date of last systems audit	.	     .
                                                                 Number of Monitors
     Pollutant                                   Added                  Deleted                  Relocated
     Sulfur.Dioxide                           	              	   .      •      	
     Nitrogen Dioxide                         	              	'                  	
     Carbon Monoxide                        	              	               	•
     Ozone
    Lead
(h) Briefly discuss changes to the Air Monitoring Network planned for the next audit period. (Discuss equipment
    needs in Section B.3.g)
(i)  Does an overall SLAMS/NAMS Monitoring Plan exist?
    Yes	 Wo	
(j)  Has the agency prepared and implemented Standard Operating Procedures for all facets of agency operation?
    Yes        No
    If no, list subject of any missing SOPs .
(k)  Do the Standard Operating Procedures adequately address at least the Woven ff# item quality control program
    required by Appendix A to 40 CFR 58?  Yes        No        Comment	
                                                   LF-4

-------
                          Section 2.0.11                30                April 1985


  (I)   Clearly identify by section number and/or document title, major changes made to documents since the last on-
      site review.


                    Title/Section #                                         Pollutants) Affected
 (m) Does the agency have an implemented plan for operations during emergency episodes? Yes       No
     Indicate latest revision, approval date and current location of this plan.
     Document Title.

     Revision Date _

     Approved	
 (n)  During episodes, are communications sufficient so that regulatory actions are based on real time data?

      Yes	  No	

 M Identify the section of the emergency episode plan where quality control procedures can be found.
 2. NETWORK DESIGN AND SITING

 (a)  Indicate By SAROAD Number any non-conformance with the requirements of 40 CFR 58, Appendices- D and E.



                                      Site ID
     Monitor	(SAROAD)	          Reason for Non-Conformance


      SO,
     03
     CO
     NO2
      PM-IO
     Pb
(b)  Please provide the following information on your previous Network Review required by 40 CFR 58.20d.

    Review performed on:    Date	

    Performed by:    .	
    Location and Title of Review Document:       •	

                                                  LF-5

-------
                            April! 986                31                Section 2.0.11

Briefly discuss all problems uncovered by this review.
(c)  Have NAMS Hard Copy Information Reports (NHCIRs) been prepared and submitted for all monitoring sites
    within the network?
    Yes         No
(d) Does each site have the required information including:
                                                   .                YES      NO
    SAROAD identification number?                                	
    Photographs/slides to the four cardinal compass points?           	
    Startup and shutdown dates?                                   	   .
    Documentation of instrumentation?                             	
    Reasons for periods of missing data?            •                	
(e) Who has custody of the current network documentation?
                           ___

(f)  Does the current level of monitoring effort, site placement, instrumentation, etc., meet requirements imposed by
    current grant conditions? Yes	No	  Comment __—	
 (g)  How often is the network design and siting reviewed?
     Date of last review.
 (h)  Please provide a summary of the monitoring activities conducted as the SLAMS/NAMS network by the agency
     as follows:
      I. Monitoring is seasonal for (indicate pollutant and month of high and low concentrations).
                                                     Month(s)
                                         High                          Low
          Pollutant                   Concentration                  Concentration                  Collocated
       .	                 	'      '	                     Y/N
       	                 	                  	        '             Y/N
       	                       '                        	                     Y/N
       	              '    	       .   ,           Y/N
       	                 	                  	                     Y/N
       	                 	                  	                     Y/N
                                                   LF-6

-------
                          Section 2.0.11                32               April 1985

      //. Monitoring is year-round for (indicate pollutant)
                                   Pollutant                          Collocated
                                                                         Y/N
                                                                         Y/N
                                                                         Y/N
                                                                         Y/N
                                                                         Y/N
 (I)  Does the number of collocated monitoring sites meet the requirements of 40 CFR 58 Appendix A?
     Yes	 No	  Comment	-	
 (j)   Does your agency monitor and/or analyze for non-criteria air and/or toxic air pollutants?  Yes	 No
     If yes, please complete the form below.
                                                Monitoring            .                       SOP Available
     Pollutant                               Method/Instrument                                   Yes/No
3. ORGANIZATION, STAFFING AND TRAINING
(a)  Please indicate the key individuals responsible for the following:
    Agency Director	,	.	
    SLAMS Network Manager
    Quality Assurance Officer .
    Field Operations Supervisor
    Laboratory Supervisor	
    Data Management Supervisor
    SLAMS Reporting Supervisor
(b)  Please indicate the number of people available to each of the following program areas:

    ,,„_,__. ..                    "                                                   Comment on Need for
    Program Area	Number	  Additional Personnel
    Network Design and Siting

    Resources and Facilities

    Data and Data Management

    QA/QC

                                                  LF-7

-------
                             April! 985                 33                Section 2.0.11
 (c)  Does the agency have an established training program?
     Yes         No
      I. Where is this documented?  „	

                (rev date)
     //. Does it make use of seminars, courses, EPA sponsored college level courses?  Yes	  No
     III. Indicate below the three (3) most recent training events and identify the personnel participating in them.
                     Event                     Dates                           Participant(s)
(d)  Does the agency subscribe to recognized publications? Please provide a list of periodicals. Are periodicals avail-
    able to all personnel?
                               Periodical Title                           Distribution
4.  FACILITIES
(a)  Identify the principal facilities where the work is performed which is related to the SLAMS/NAMS network? (Do
    not include monitoring sites but do include any work which is performed by contract or other arrangementsT~
                    Facility                Location                Main  SLAMS/NAMS Function
                                                    LF-8

-------
                          Section 2.0.11
                                                      34
                                                                        April 1986
 (b)  Please review the entries on the above table. Ara there any areas of facilities which you believe should be up-
     graded? Please identify by location.
 (c) Are there any significant changes which are likely to be Implemented to agency facilities before the next systems
    audit? Comment on your agency's needs for additional physical space (laboratory, office, storage, etc.)
                     Facility
Function
Proposed Changs - Date
                                            B. FIELD OPERATIONS

1.  ROUTINE OPERATIONS

(a)  Is the documentation of Monitoring Standard Operating Procedures complete?

    Yes	  No	

    Please complete the table below.
           Pollutant
           Monitored
                                                         Date of Last Revision
          PM-IO
     Pb
     SO,
     NO,
                (continuous)
                  {bubbler*}
    CO
    Others (list by pollutants)
                                                   LF-9

-------
                             April 1985                35                 Section Z.O.I 1
(b)  Are such procedures available to all field operations personnel?
    Yes         No         Comment	
(c)  Are standard operating procedures prepared and available to field personnel which detail operations during
    episode monitoring?
    Yes         No         Comment
(d)  For what does each reporting oganization within the agency monitor? Provide the list requested below.
    Reporting Organization             •         # of Sites                      Pollutants
(e)  On the average, how often are most of your sites visited by a field operator?  	per
(f)  Is this visit frequency consistent for all reporting organizations within your agency? Yes       No
    If no, document exceptions     •	           •  •	'.	
(g)  On the average, how many sites does a single site operator have responsibility for?.
(h)  How many of the sites of your SLAMS/NAMS network are equipped with manifold(s) #
    I. Briefly describe most common manifold type..	
   //. Are manifolds cleaned periodically?  Yes	No
      If yes, how often?  	per	
   ///. If the manifold is cleaned, what is used?.
   IV.  Are manifold(s) equipped with a blower?   Yes	No '	
    V.  Is there sufficient air flow through the manifold at all times?
       Yes	Wo	
       Approximate air flow is
                                      (flow units}
   VI. Is there a conditioning period for the manifold after cleaning? Briefly comment on the length of time the condition-
      ing is performed.	
                                                   LF-10

-------
                         Section 2.0.11                36                April 1985


 (!)   What material Is used for instrument lines?	
 (j)   Has the agency obtained necessary waiver provisions to operate equipment which does not meet the effective
     reference and equivalency requirements?   Yes	No	
     Comment on Agency use of approved/non-approved instrumentation..
(k)  Please complete the table below to indicate which analyzers do not conform with the requirements of 40 CFR 53
    for NAMS, SLAMS, or SIP related SPM's.

                                                                      Site                  Comment on
 Pollutant             Number            • Make/Model             Identification              Variances


  CO
   NO2



   03
(I)  Please comment briefly and prioritize your currently identified instrument needsT
 2. QUALITY CONTROL

 (a)  Are field calibration procedures included in the documented Standard Operating Procedures? Yes
     No
     Comment on location (site, lab, office) of such procedures.
(b)  Are multipoint calibrations performed? Indicate both the frequency and pollutant.

    Reporting Organization                  Pollutant                  Frequency
(c)  Are calibrations performed in keeping with the guidance offered in Section 2.0.9 Vol. II of the Quality Assurance
    Handbook for Air Pollution Measurement Systems?  Yes	N.o                                                   ^fcl
                      	           '                     •                              w
If no, why not?.
                                                  LF-11

-------
                               April 1985                37                Section 2.0.11

   (d)  Are calibration procedures consistent with the operational requirements of Appendices to 40 CFR 50 or to ana-
       lyzer operation/instruction manuals?  Yes       No
       If no, briefly explain deviations.
   (e)  Have changes been made to calibration methods based on manufacturer's suggestions for a particular instru-
       ment?  Yes ___^ No	
       Are these also documented?  Yes •No
   (f)  ^Jh"*dfrd?Ate"al5- US*Bd f?r calibrations meet tne requirements of appendices to 40 CFR 50 (EPA. reference
       methods} and Appendix A to 40 CFR 58 (traceability of materials to NBS-SRMs or CRMs)?  Yes.	No	
       Comment on deviations	;	
  (g)  Are all flow-measurement devices checked and certified?
      Yes	No        Comment	
  (h)  What are the authoritative standards used for each type of flow measurement? Please list them in the table be-
      low, indicate the frequency of calibration standards to maintain field material/device credibility.
      Flow Devices      •            Primary Standard                  Frequency of Calibration
 (i)   Where do field operations personnel obtain gaseous standards?
     Are those standards certified by:                                                         Y            v
       The agency laboratory?
       EPA/EMSL/RTP standards laboratory?
       A laboratory separate from this agency but part of the same reporting organization?       	       	
       The vendor?
       NBS?
 (j)  Does the documentation include expiration date of certification? • Yes     '  No
     Reference to primary standard used?  Yes       No
     What traceability protocol is used? ___	
     Please attach an example of recent documentation of traceability (tag, label, log sheet).
}(k)  Is calibration equipment maintained at each site?  Yes       No
     For what pollutants?	___
                                                   LF-12

-------
                         Section 2.0.11                38                April 1985



(II  How is the functional integrity of this equipment documented?	
(m) Please complete the table below for your agency's site standards (up to 7% of the sites, not to exceed 20 sites).


                                   Primary                     Secondary                     Recertification
    Parameter                    Standard                      Standard                          Date


      CO



      NO2



      S02
(n)  Are level. 1 zero and span (z/s) calibrations (or calibration checks) made for all continuous monitoring equipment
    and flow checks made for TSP samplers?  Yes	No  	
    Please complete table below:
                                                     Span Cone.
                             Pollutant                   (ppm)                   Frequency
    L Continuous analyzers
                             Flow Rate                       Frequency
            O                	                       —a	-
    II,  7W Samplers          	       .                	
(o)  Does the agency have acceptance criteria for zero/span checks?  Yes       No        Comment.
     I. Are these criteria known to the field operations personnel?

       Yes	No	

    //. Are they documented in standard operating procedures?

       Yes	No	

                                                  LF-13

-------
                              April 1986                39                Sactfon 2.0.11

        If not, indicate document and section where they can be found.


     '"'       3 d°YeUsmentS di^ussedI" W above indicate when zero/span adjustments should and should not be
        Indicate an example	. 	^	
     IV. Are zero and span check control charts maintained?  Yes       No

 (P>  checks?"9 Wlth 4° °FR 5S reaulations' are a"Y neeasaary zero and span adjustments made after precision
     Yes	No	
    . If no, comment on why not.
 (q)  Are precision check control charts maintained?  Yes       No
 (r)   Who has the responsibility for performing zero/span checks?
                                                                                               ' meets
    Please comment on any discrepancies.	
 W  Please Fdemify Person(S) .wftn the responsibly for performance of precision checks on continuous analyzers.
     Person(s)	    .	                                                           .  -
     Title .
3.  PREVENTIVE MAINTENANCE

                                      • special training in performing preventive maintenance? Briefly comment
ib)  Is this training routinely reinforced?  Yes -      No
    If no, why not? J	
                         ^oZ^ZXSSSSZ'li ±e%r !nStrUmem ^f^ers? ^cate be,OW or
                                                  UF-14

-------
                         Section 2.0.11                40                April 1985

(f)  Comment briefly on the adequacy and availability of the supply of spare parts, tools and manuals available to the
    field operator to perform any necessary maintenance activities. Do you feel that this is adequate to prevent any
    significant data loss?	—	———
(g) Is the agency currently experiencing any recurring problem with equipment or manufacturer(s)? If so, please
    identify the equipment and/or manufacturer, and comment on steps taken to remedy the problem.
4, ^RECORD KEEPING
(a)  Is a log book(s) maintained at each site to document site visits, preventive maintenance and resolution of site
    operational problems and corrective actions taken?  Yes	No	  Other uses	;—:	
(b) Is the logbook maintained currently and reviewed periodically?
    Yes	No	 Frequency of Review	
(c)   Once entries are made and all pages filled, is the logbook sent to the laboratory for.archiving?
     Yes	No	                                                  •
     // no, is it stored at other location(s) (specify)
(d)  What other records are used?               '       '                     YES      NO
    Zero/span record?                        .
    Gas usage log?             '
    Maintenance log?
    Log of precision checks?
    Control chans?
    A record of audits?     .                          '
    Please describe the use and storage of these documents.
(e)  Are calibration records or at least calibration constants available to field operators? Yes	No         Please
    attach an example field calibration record sheet to this questionnaire.
                                                   LF-15

-------
                             April 1885                41                Section 2.0.11


 5. DATA ACQUISITION AND HANDLING

 (a)  With the exception ofTSP are instrument outputs (that is data) recorded to (a) stripcharts, (b) magnetic tape
     acqu,s,t,on system (c) digit,zed and telemetered directly to agency headquarters? Please complete the table
     below for each of the reporting organizations, or agencies within the overall P.O.


     Beportiny Organization                     Pollutants
(b)  Is there stripchart backup for all continuous analyzers?  Yes       No

(c)  Where is the flow of high-volume samplers recorded at the site?

    For samplers with flow controllers?  Log sheet	, Dixon chart	, Other	(specify)-

    On High-volume samplers without flow controllers?  Log sheet	, Dixon chart	, Other	(specify)


                                                      equ'pment are available to th*field


-------
                         Section 2.0.11
42
April 1986
                                        C. LABORATORY OPERATIONS


1. ROUTINE OPERATIONS


(a)  What analytical methods are employed in support of your air monitoring network?
                       Analysis
                      Methods
      Pb




      SO4



      NO3




      SO2




      NO2
      Others (list by pollutant)
(b)  Are bubblers used for any criteria pollutants in any agencies?  Yes
                  No
    indicates the number of sites where bubblers are used, the agency and pollutantfs}.
           If yes, attach a table which
(c)  Do any laboratory procedures deviate from the reference, equivalent, or approved methods? Yes	No	
    If yes, are the deviations for lead analysis      , fSft filter conditioning       or other	(specify below)?
                                          	
(d)  Have the procedures and/or any changes been approved by EPA?  Yes

    Date of Approval  	
                      No
                                                  LF-17

-------
                            April 1985                43                Section 2.O.1 1





(e)  Is the documentation of Laboratory Standard Operating Procedures complete?  Yes       No        Please

    complete the table below.                                                      -    -





                        Analysis                                             Method
     •f&ft.
      Pb
      S04
      N03
     SO,
     NO2
     Others (list by pollutant)
(f)  I? sufficient instrumentation available to conduct your laboratory analyses?  Yes       No     '   If no ol ease
    indicate instrumentatio                                  '      -            -    - '     ' please
   indicate instrumentation needs in the table below.
           Instrument                                          */,,,.. -.                     ..
             ...   .  ,                                            /VGVv Or                     Yf*£)r nf

             Needed _ Analysis                 Replacement                AcquisMon
                                                 LF-18

-------
                        Section 2.0.11                44                April 1985

2,  QUALITY CONTROL
(a)  Please complete the table for your agency's laboratory standards.

                                      Primary                Secondary               decertification
    Parameter                        Standard                Standard                    Date
    CO
    NO2
    SO2
    Weights
    Temperature
    Moisture
    Barometric Pressure
    Flow
    Lead
    Sulfate
    Nitrate
(b)  Are all chemicals and solutions clearly marked with an indication of shelf life?  Yes	No
(c)  Are chemicals removed and properly disposed of when shelf life expires?  Yes	No	
(d)  Are only ACS chemicals used by the laboratory?  Yes	No	
(e)  Comment on the traceability of chemicals used in the preparation of calibration standards—
                                                  IF-19

-------
                             April 1985                45                Section 2.0.11

 (f)   Does the laboratory:
     Purchase standard solutions such as those for use with lead or other AA analysis?  Yes	No	Make them
     themselves?  Yes	No	If the laboratory staff routinely make their own standard solutions, are procedures
     for such available?  Yes	No	Where?— Attach an example.
 (g)  Are all calibration procedures documented?  Yes	No	
     Where?  	    	    ___^_^_^__
                     (title)                 (revision)
     Unless fully documented, attach a brief description of a calibration procedure.
 (h)  Are at least one duplicate, one blank, and one standard or spike included with a given analytical batch? Yes
     	No	Identify analyses for which this is routine operation.
(i)   Briefly describe the laboratory's use of data derived from blank analyses.
    Do criteria exist which determine acceptable/non-acceptable blank data? 'Please complete the table below.
                Pollutant              '                       Blank Acceptance Criteria
                  SO2                    .      _ _ _ _ -
                 • NO2           '          '  •         • _ _ _ ; _ ; _ •
                  SO4                          __ _
                  N03   .                       _
                  Pb
                  Other
(j)  How frequently and at what concentration ranges does the lab perform duplicate analysis? What constitutes ac-
    ceptable agreement? Please complete the table -below.
                Pollutant                        Frequency                    Acceptance Criteria
                  S02
                '  NO2                           _ __ _ '
                  SO4
                  NO3                           _ _ _ _
                  Pb
                  Other
                                                   LF-20

-------
                         Section 2.0.11               .46                April 1985

 (k)  How does the lab use data from spiked samples? Please indicate what may be considered acceptable percentage
     recovery by analysis? Please complete the table below.
                 Pollutant                                 % Recovery Acceptance Criteria
                  SO2  Bubblers                	
                  NO2  Bubblers	
                  SO4                          	,	
                  N03	
                  Pb	
                  TSP                              •	!	-
                  voc            .	;	
                  Other	:	
 (I)  Does the laboratory routinely include samples of reference material obtained from EPA within an analytical
    batch? Yes       No
    If yes, indicate frequency, level, and material used.
(m) Are mid-range standards included in analytical batches? Yes  	No	If yes, are such standards included
    as a QC check (span check) on analytical stability? Please indicate the frequency, level and compound used in the
    space provided below.	
(n) Do criteria exist for "real time" quality control based on the results obtained for the mid-range standards dis-
    cussed above? Yes	 No	If yes, briefly discuss them below or indicate the document In which they
    can be found.	
(o)  Are appropriate acceptance criteria documented for each type of analysis conducted? Yes	No       Are
    they known to at least the analysts working with respective instruments?  Yes	A/o             	
3, PREVENTIVE MAINTENANCE
(a)  For laboratory equipment, who has responsibility for major and/or minor preventive maintenance?
    Person	.	    Title	
(b)  Is most maintenance performed: in the lab? Yes	No 	in the instrument repair facility? Yes
    No	at the manufacturer's facility? Yes	No
(c)  Is a maintenance log maintained for each major laboratory instrument?
    Yes       No        Comment	
                                                  LF-21

-------
                             April 1986                47                Section 2.0.11

 (d)  Are service contracts in place for the following analytical instruments:
                                                                          YES      NO
     Analytical Balance                                                    	    	
     Atomic Absorption Spectrometer                                     '  	    	
     Ion Chromatograph                                                   	    	
     Automated Colorimeter                                               	    	
 4.  RECORD KEEPING
 (a)  Are all samples that are received by the. laboratory: logged- in? Yes	No 	 assigned a unique labora-
     tory sample number? Yes	No	 routed to the appropriate analytical section? Yes	No	
     Discuss sample routing and special needs for analysis (or attach a copy of the latest SOP which covers this).
     Attach a flow chart if possible.
 (b)  Are logbooks kept for all analytical laboratory-instruments? Yes	Wo	
 (c)  Do these logbooks indicate:
                                                                          YES      NO
    , analytical batches processed?                                    •      	:    	
     quality control -data?.                                            •     	    	
     calibration data?                                                     '.	    	
     results of blanks, spikes and duplicates?                                	    	
     initials of analyst?                                                    	    	
 (d)  Is there a logbook which indicates the checks made on: weights? Yes  	No _^_ humidity indicators?
     Yes	No	balances? Yes	No	thermometer(s)? Yes       No
 (e)  Are logbooks maintained to track the preparation of filters for the field? Yes	No	Are they current?
     Yes	No __^__ Do they indicate proper use of conditioning?  Yes        No       Weighings? Yes
     No	Stamping and numbering? Yes	No	
 (f)  Are logbooks kept which track filters returning from the field for analysis? Yes	No	
 (gi  How are data records from the laboratory archived?
     Where?	
     Who has the responsibility? Person.
     Title	
     How long are records kept? Years .
 (h)  Does a chain-of-custody procedure exist for laboratory samples? Yes       No
 (i)   Has chain-of-custody been documented and implemented as part of standard laboratory procedures? Yes
k    No	If yes, indicate date, title and revision number where it can be found.
                                                   UF-22

-------
                          Section 2.0.11
                                                      48
        April 1985
 5. DATA ACQUISITION AND HANDLING

 (a)  Identify those laboratory instruments which make use of computer interfaces directly to record data. Which ones
     use stripcharts? integrators?
 (b)  Are QC data readily available to the analyst during a given analytical run? Yes	No	

 (c)  For those instruments which are computer interfaced, indicate which are backed up by stripcharts?.
(d)  What is the laboratory's capability with regard to data recovery? In case of problems, can they recapture data or
     are they dependent on computer operations? Discuss briefly.
(e) Has a user's'manual been prepared for the automated data acquisition instrumentation? Yes	No
    Comment	                     '
    Is it in the analyst's or user's possession? Yes	No	   Is it current? Yes
                     No
(f)  Please provide below a data flow diagram which establishes, by a short summary flow chart: transcriptions, vali-
    dations, and reporting format changes the data goes through before being, released to the data management  •
    group. Attach additional pages as necessary.
6. SPECIFIC POLLUTANTS: T3f* AND LEAD

    f§£ PH-IO

(a) Are filters supplied by EPA used at SLAMS sites? Yes	/Vo

    Comment	'.	
(b)  Do filters meet the specifications in the Federal Register 40 CFR 50? Yes       No       Comment
(c)  Are filters checked for surface alkalinity? Yes	No

    Indicate frequency	
Id)  Are filters visually inspected via strong light from a view box for pinholes and other imperfections? Yes
    No	// no, comment on way imperfections are determined?	'
(e)  Are filters permanently marked with a serial number? Yes
    plished:.	     '
No
Indicate when and how this is accom-
(f)  Are unexposed filters equilibrated in controlled conditioning environment which meets or exceeds the reauire-
    ments of 40 CFR 50? Yes	No	If no, why not?
                                                   LF-23

-------
                             April 1985                49                Section 2.O.11
 (g)  Is the conditioning environment monitored? Yes	No
     Indicate frequency	
     Are the monitors properly calibrated? Yes	No
     Indicate frequency	
 (h)  Is the balance checked with Class "S" weights each day it is used? Yes	No	If no, indicate frequency
     of such checks
 (i)  Is the balance check information placed in QC logbook? Yes	No
     If no, where is it recorded?	'.	
 (j)  Is the filter weighed to the nearest milligram? Yes	No	If not, what mass increment.
 (k)  Are filter serial numbers and tare weights permanently recorded in a bound notebook? Yes	No       If
     no, indicate where	^___ZIIII_ZZZZ__
 (I)  Are filters packaged for protection while transporting to and from the monitoring sites? Yes       No
 (m)  How often are filter samples collected? (Indicate average lapse time (hrs.) between end of sampling and labora-
     tory receipt.)
(n) Are field measurements recorded in logbook or on filter folder?      •

(o) Are exposed filters reconditioned for at least 24 hrs in the same conditioning environment as for unexposed
    filters? Yes       No                         .
    If no, why not?.
(p)  Are exposed filters removed from folders, etc., before conditioning? Yes       No
(q)  Is the exposed filter weighed to the nearest milligram? Yes	No
(r)  Are exposed filters archived? Yes	No       When?	
    Where?	:	
    Indicate retention period.
(s)  Are blank filters reweighed? Yes	No	If no, explain why not.
    If yes, how frequently?.
                                              N°
                                                   LF-24
                                                                analyses °ther than Pb and ™** which are

-------
                          Section 2.0.11                 60                ApriM985


 (u)  Are sample weights and collection data recorded in a bound laboratory logbook? Yes	No	On data
     forms? Yes	No	

 M  Are measured air volumes corrected to reference conditions as given in CFR regulations (Q3ta of 760 mm Hg and
     25"C) prior to calculating the Pb concentration? Yes	No	
     If not, indicate conditions routinely employed for both internal and external reporting.
     LEAD

 (a)  Is analysis for lead being conducted using atomic absorption spectrometry with air acetylene flame?
     Yes       No
     If not, has the agency received an equivalency designation of their procedure?.
 (b)  Is either the hot acid or ultrasonic extraction procedure being followed precisely? Yes       No       Which?
 (c)  Is Class A borosilicate glassware used throughout the analysis? Yes	  No
(d)  Is all glassware scrupulously cleaned with detergent, soaked and rinsed three times with distilled-deionized
     water? Yes	No	If not, briefly describe or attach procedure.
M If extracted samples are stored, are linear polyethylene bottles used? Yes        No       Comment.
(I)  Are all batches of glass fiber filters tested for background lead content? Yes       No       At a rate of 20 to
    30 random filters per batch of 500 or greater? Yes	No	Indicate Tale      ~~



tg)  Are ACS reagent grade HNO3 and HCI used in the analysis? Yes	No	If not, indicate grade used
(h)  Is a calibration curve available having concentrations that cover the linear absorption range of the atomic absorp-
    tion instrumentation? Yes	No	Briefly describe	
(1)  Is the stability of the calibration curve checked by alternately remeasuring every 10th sample a concentration
    S1 \ng Pb/ml; &10 n<7 Pb/ml? Yes	No	If not, indicate frequency.
                                                   LF-25

-------
                             April 1985                51                 Section 2.0.11


 (j)  Are measured air volumes corrected to reference conditions as given in CFR regulations (Qstd of 760 mm Hg and
    25°C) prior to calculating the Pb concentration? Yes	No	If not, indicate conditions routinely em-
    ployed for both internal and external reporting.
(k) •  In either the hot or ultrasonic extraction procedure, is there always a 30-min H^O soaking period to allow
     trapped in the filter to diffuse into the rinse water? Yes       No       Comment
(I)   Is a quality control program in effect that includes periodic quantification of (1) lead in 3/4" x 8" glass fiber filter
     strips containing 100-300 |xg Pb/strip, and/or (2) a similar strip with 600-1000 jig strip, and (3) blank filter strips
     with zero Pb content to determine if the method, as being used, has any bias? Yes	No	Comment on
     lead QC program or attach applicable SOP.	ZZZI__	-
(m) Are blank Pb values subtracted from Pb samples assayed? Yes	No ^_^ If not, explain why.
                                       D. DATA AND DATA MANAGEMENT

1.  DATA HANDLING

(a)  Is there a procedure, description, or a chart which shows a complete data sequence from point of acquisition to
    point of submission of data to EPA?  Yes	  No	

    Please provide below .a data flow diagram indicating both the data flow within the reporting organization and the
    data received from the various local agencies.
(b)  Are data handling and data reduction procedures documented?

    For data from continuous analyzers?  Yes	  No	

    For data from non-continuous methods?   Yes        No
(c)  In what format and medium are data submitted to data processing section? Please provide separate entry for
    each reporting organization.

                    Reporting Organization                Data Medium                Format
(d)  How often are data received at the processing center from the field sites and laboratory?  at least once a
    week?	  every 7- 2 weeks?	  once a month?	

(e)  Is there documentation accompanying the data regarding any media changes, transcriptions, and/or flags which
    have been placed into the data before data are released to agency internal data processing?  Describe.
                                                    LF-26

-------
                          Section 2.0.11                52                April 1985



 (f)  How are the data actually entered to the computer system? Digitization of stripcharts? Manual or computerized
     transcriptions? Other?
 (g)  Is a double-key entry system used for data at the processing center?  Yes	No	Are duplicate card decks
     prepared?  Yes	No	/f no, why not?
 (hi  Have special data handling procedures been adopted for air pollution episodes? Yes        No       'If yes,
     provide brief description.                                              .
 2. SOFTWARE DOCUMENTATION


 (a)  Does the agency have available a copy of the AEROS Manual? Yes        No        Comment
 (b)  Does the agency have the PARS user's guide available? Yes   •     No __	._ Comment (provide guide #)
(c)  Does the Data Management Section have complete software documentation? Yes        No

    Comment	
    If yes, indicate the implementation date and latest revision dates for such documentation.
(d)  Do the documentation standards follow the guidance offered by the EPA Software Documentation Protocols?
    Yes        No
    If no, what protocols are they based on?.
lei  What is the origin of the software used to process air monitoring data prior to its release into the SAROAD/NADB
    database?
     I. Purchased? Yes	  No	; Supplier.

       Date of latest version	
                                                 LF-27

-------
                             April 1985                53                Section 2.0.11
     //. Written in-house? Yes	  No	/ Latest version.

        Date	
     ///. Purchased with modifications in-house?  Yes        No
        Latest version	  Date	

     IV. Other (specify)	

 (f)   Is a user's manual available to data management personnel for all software currently in use at the agency for
     processing SLAMS/NAMS data?
     Yes        No        Comment.
 (g)  Is there a functional description either: included in the user's manual?  Yes	  No	 separate from it
     and available to the users? Yes	  No '_	

 (h)  Are the computer system contents, including ambient air monitoring data backed up regularly?   Yes	 Nn
     Briefly describe, indicating at least the media, frequency, and backup-media storage location	
(i)   What is the recovery capability (how much time and data would be lost) in the event of a significant computer
     problem?	
(j)  Are test data available to evaluate the integrity of the software?  Yes	Nn     Is it 'properly documented?
    Yes	~~No	                       .           .

3.  DATA VALIDATION AND CORRECTION                                                               .    •

(a)  Have validation criteria, applicable to all pollutant data processed by the reporting organization been established
    and documented?  Yes	  No

    If yes, indicate document where such criteria can be found (title, revision date).	
(b)  Does documentation exist on the Identification and applicability of flags (i.e., identification of suspect values)
    within the data as recorded with the data in the computer files? Yes         No,


(c)  Do documented data validation criteria employ address limits on and for the following:

     1. Operational parameters, such as flow rate measurements or flow rate changes.	
     //. Calibration raw data, calibration validation and calibration equipment tests.
    III. All special checks unique to a measurement system .
    IV. Tests for outliers in routine data as part of screening process .
                                                   LF-28

-------
                           Section 2.0.11                 54                April 1985



      V, Manual checks such as hand calculation of concentrations and their comparison with computer-calculated data
 (d) Are changes to data submitted to NADB documented in a permanent file? Yes	  No        If no, why
     not?	
 (e)  Are changes performed according to a documented Standard Operating Procedure or your Agency Qualitv Assur-
     ance Project Plan? Yes	  No-	

     If not according to the QA Project Plan, please attach a copy of your current Standard Operating Procedure.

 (f)  Who has signature authority for approving corrections?
                        (Name)                                        (Program Function)

                        ummaries prepared at each critical point in tl
                        (he applicable block of data to the next level <

     Please indicate the points where such summaries are performed.
(g)  Are data validation summaries prepared at each critical point in the measurement process or information flow
    and forwarded with the applicable block of data to the next level of validation? Yes         No
 (h)  What criteria are applied for data to be deleted? Discuss briefly.
 (i)   What criteria are applied to cause data to be reprocessed? Discuss.
(j)  Is the group supplying data provided an opportunity to review data and correct erroneous entries?
    Yes	  No	 If yes, how?
    Are corrected data resubmitted to the issuing group for cross-checking prior to release? Yes        No
4,  DATA PROCESSING

(a)  Does the agency generate data summary reports? Yes         No
    Are the data used for in-house distribution and use? Yes         No

    Publication? Yes	  No

    Other (specify)	
                                                   LF-29

-------
                            April 1985               BS                Section 2.0.11

(b) Please list at least three (3) reports routinely generated, providing the information requested below.
              Report Title                         Distribution                 •   Period Covered
(c)  Have special procedures been instituted for pollution index reporting? Yes	  No	 // yes, provide
    brief description.                  .           .             •                .
(d) Who at the agency has the responsibility for submitting data to SAROAD/NADB? (name) _	
    (title)	!	
    Is the data reviewed and approved by an officer of the agency prior to submittal?  Yes	  No
    (name)	;	;	  (title)	!,	
(e)  Are those persons different from the individuals who sumbit data to PARS? Yes	  No        If yes,
    provide name and title of individual responsible for PARS data submittal.
    (name)	.	  (title)      '      	,	  PARS
  '  Data review and approval (name)	:	
 ,   (title)	,	r_^___	
(f)  How often are data submitted to:     ~
    SAROAD?	:	;	._
    PARS?	:	
(g)  How and/or in what form are data submitted?
    TO SAROAD?	!	
    TO PARS?	
(h)  Are the recommendations and requirements for data coding and submittal, In the AEROS User's Manual followed
    closely for SAROAD?  Yes	  No	Comment on any routine deviations in coding procedures.
(I)  Are the recommendations and requirements for data 'coding and submittal, in the PARS User's Guide, followed
    closely?  Yes	  No	 Comment on any routine deviations in coding and/or computational'procedures.
(j)   Does the agency routinely request a hard copy printback on submitted data:
    from SAROAD/NADB?    Yes	  No	
    from PARS?             Yes        No
                                                 LF-30

-------
                          Section 2.0.11                66                 April 1985

 fk)  Are records kept for at least 3 years by the agency in an orderly, accessible form? Yes	  No
     If yes, does this include raw data	, calculation	, QC data	, and reports	? If no, please
     comment.                                                                        	
 (I)   In what format are data received at the data processing center? (Specify appropriate pollutant.)
     (a) concentration units __.	(b) % chart	(c) voltages	(d) other	
 (m)  Do field data include the following documentation?
   _ Site ID? Yes	  No	
     Pollutant type?  Yes	  No
    Date received at the center? Yes	  No	
    Collection data (flow, time date)? Yes	  No	
    Date of Laboratory Analysis (if applicable)  Yes	  No
    Operator/Analyst?  Yes	  No	
(n) Are the appropriate calibration equations submitted with the data to the processing center?  Yes	No
    If not, explain.
(o)  Provide a brief description of the procedures and appropriate formulae used to convert field data to concentra-
    tions prior to input into the data bank.
    SO,
    N02.
    CO.
    03.
    TSP.
    cwyrwc.
   Pb.
   Other.
                                                   LF-31

-------
                             April 1985                57                Section 2.0.11



 (p)  Are all concentrations corrected to EPA standard (298°K, 760 mm Hg) temperature and pressure condition before
     input to the SAROAD?
     Yes	  No	 If no, specify conditions used.
 (q)  Are data reduction audits performed on a routine basis? Yes	  No	  If yes,

     at what frequency?-	
     are they done by an independent group?.
 (r)  Are there special procedures available for handling and processing precision, accuracy, calibrations and span
     checks? Yes        No
     If no, comment.
     If yes, provide a brief description: Span check data,

     Calibration data	'.	

     Precision data	

     Accuracy data	  •
 (s)  Are precision and accuracy data checked each time they are recorded, calculated or transcribed to ensure that
     incorrect values are not submitted to EPA? Yes	 No	 Please comment and/or provide a brief de-
     scription of checks performed.	
 ftj  /s a final data processing check performed prior to submission of any data?  Yes.	No.

    If yes, document procedure briefly	:	__	'
    If no, explain.
5.  INTERNAL REPORTING


(a)  What reports are prepared and submitted as a result of the audits required under 40 CFR -Appendix A?

                 Report Title         .                           .                  . Frequency
    (Please include an example audit report and, by attaching a coversheet, identify the distribution such reports are
    given within the agency.)


(b)  What internal reports are prepared and submitted as a result of precision checks also required under 40 CFR 58
    Appendix A?


                 RePQrt                                                           Frequency
    (Please include an example of a precision check report and, identify the distribution such reports receive within
    the agency.)
                                                   LF-32

-------
                        Sactlon 2.0,11                 B8                April 1986


 (c)  Do either the audit or precision reports Indicated include a discussion of corrective actions initiated based on
     audit or precision results?

     Yes	  No	  // yes, identify report(s) and section numbers.
 (d)  Does the agency prepare Precision and Accuracy summaries other than Form  1? Yes	  No	 If no,
     please attach examples of recent summaries including a recent Form 7.

 (e)  Who has the responsibility for the calculation and preparation of data summaries? To whom are such P and A
     summaries delivered?

             Name                       Title       .            Type of Report                 Recipient
 (f)   Identify the individual within the agency who receives the results of the agency's participation in the NPAP and
     the internal distribution of the results once received.

     Principal Contact.for NPAP is (name, title)
     Distribution is.
                                  (name)            •                    (title)

6. EXTERNAL REPORTING

(a)  For the current calendar year or portion thereof which ended at least 135 calendar days prior to the receipt of this
     questionnaire, please provide the following percentages for required data submitted.

                                              %Submitted on Time*


      Monitoring Qtr.            SO2         CO    '     O3          NO2          W         Pb
     1 (Jan. 1-March 31)
    2 (Apr. 1-June 30)
    3 (July 1-Sept. 30)
    4 (Oct. 1-Dec. 31)
    '"On-Time" - within 135 calendar days after the end of the quarter in which the data were collected.

(b)  Identify the individual within the agency with the responsibility for preparing the required 40 CFR 58 Appendix F
    and G reporting inputs.
    Name		  Title .

                                                         LF-33

-------
                           April 1985
                                                     59
                                                                       Section 2.O.11
 (c)  Identify the individual within the agency with the responsibility for reviewing and releasing the data.

     Name	:	:	  Title	
 (d)  Does the agency regularly report the Pollutant Standard Index (PSI)? Briefly describe the media, coverage, and
     frequency of such  reporting.
 (e)  What fraction of the SLAMS sites (by pollutant) reported less than 75% of the data (adjusted for seasonal moni-
     toring and site start-ups and terminations)?


                                                  FY

Pollutant
Percent of Sites
<75% Data Recovery
1st Quarter 2nd Quarter 3rd Quarter 4th Quarter
Ozone • • .
Nitrogen Dioxide
Sulfur Dioxide
Carbon Monoxide
PW-io
T-— *„/ C* 	 ,«W«W D«^*,*«. /„*


Lead
(f)  Does the agency's annual report (as required in 40 CFR 58.26) include 'the following?




    Data summary required in Appendix F.                                  '


    Annual precision and accuracy information described in Section 5.2 of Appendix A.


    Location, date, pollution source and duration of all episodes reaching the significant
    harm levels.


    Certification by a senior officer in the State or his designee.


!gl  Please provide the dates at which the annual reports have been submitted for the last 2 years.
.  YES     NO
                                                    LF-34

-------
                       Section 2.0.11                60                April 1985



                                   £ ClUAUTY ASSURANCE/QUALITY CONTROL

 1. STATUS OF QUALITY ASSURANCE PROGRAM


 (a)  Does the agency have-an EPA-approved quality assurance program plan? Yes	  No

     If yes, have changes to the plan been approved by the EPA? Yes	  No

     Please provide: Date of Original Approval	       Date of Last Revision

     Date of Latest Approval	

 (b)  Do you have any revisions to your QA Program Plan still pending?  Yes	  No	

 (c)  Is the QA Plan fully implemented? Yes	  No       Comment:	
 (d)  Are copies of QA Pla.n or pertinent sections available to agency personnel?  Yes	 No.	  If no, why
     not?                                                          •.
 (e)  Which individuals routinely receive updates to QA Plan?
2. AUDITS AND AUDIT SYSTEM TRACEABILITY


(a) Does the agency maintain a separate audit/calibration support facility laboratory: Yes        No

(b) Has the.agency documented and implemented specific audit procedures? Yes        No

(c) Have audit procedures been prepared in keeping with the requirements of Appendix A to 40 CFR 58?
    Yes       No
    If no, comment on any EPA approved deviations.
(d)  Do the procedures meet the specific requirements for independent standards and the suggestions regard/no per-
    sonnel and equipment? Yes	  No       Comment:		
M  Are SRM or CRM materials used to routinely certify audit materials? Yes        No

(f)  Does the agency routinely use NBS-SRM or CRM materials?  Yes	  No	For audits only?       For
    calibrations only?	For both?	For neither, secondary standards are employed         '  	
                                                 LF-35

-------
                             April 1985                61                Section2.0.11


. (g)  Please list below areas routinely covered by this review, the date of the last review, and changes made as a direct
     result of the review.



     Pollutants                                    Audit Method                 Audit Standard


       CO



       03



       NO2     '        '      '

                     (continuous)

       SO2
       PM-/O
(h) Are SRM or CRM materials used to establish traceability of calibration and zero/span check materials provided to
   . field operations personnel? Yes _  No _

(!)  Specifically for gaseous standards, how is the traceability of audit system standard materials established? Are
    they: purchased certified by the vendor? _ __ _ ___ _ . _
    certified by the QA support laboratory which is part of this agency?
    Other? (Please comment briefly below)..
(j)  Are all agency traceability and standardization methods used documented?  Yes   -     No	 Indicate docu-
    ment where such methods can be found.                           •



(k)  Do the. traceability and standardization methods conform with the guidance of Section 2.0.7 Vol. II of the Hand-
    book for Air Pollution Measurement Systems?

    For permeation devices?  Yes	  No	       For cylinder gases? Yes	  No
(I)  Does the agency have identifiable auditing equipment (specifically intended for sole use) for audits?

    Yes	  No	 If yes, provide specific identification



(m) How often is auditing equipment certified for accuracy against standards and equipment of higher authority?.
                                                   LF-36

-------
                          Section 2.0.11                62                April! 985

     ^if! *!]? afudit^u!Pment^ecks performed, have pass/fail (acceptance criteria) been decided for this
     eqwpment? Ind.cate what these cntena are with respect to each pollutant. Where are such criteria documented?
                    Pollutant                   '                                     Criteria
 3.  NATIONAL PERFORMANCE AUDIT PROGRAM (NPAP) AND ADDITIONAL AUDITS
 M  Prl^mhe'ndMdUal Wlth P"marV responsibilitY for the reW'r*d participation in the National Performance Audit
     For gaseous materials? (name, title) __	
     For laboratory materials? (name, title).
    If yes, has the agency included QA requirements with this agreement?  Yes         No
    Is the agency adequately familiar with their QA program? Yes        No
(c)  Date last systems audit was conducted:	
    By whom?—	
(d)  Please complete the table below
    Parameter Audited                                  Date of Last NPAp
    SO,
    CO
   Pb
   flerOjiA.s
    Oy

   NO,
                                                LF-37

-------
                             Apr/11986                 63                Section Z.O.It


(e)  Does the agency participate in the National Performance Audit Program (NPAP) as required under 40 CFR 58
    Appendix A?  Yes	  No	

    If no, why not? Summarize below.
4. DOCUMENTATION AND DATA PROCESSING REVIEW

(a)  Does the agency periodically review its record-keeping activities?  Yes	  No
    Please list below areas routinely coverd by this review, the date of the last review, and changes made as a direct
    result of the review.

    Area/Function             Date of Review             Changes?             Discuss Changes

    :	             	                Y/N                	

    	            '	                Y/N                	

    	             	                Y/N	

(b)  Are data audits (specific re-reductions of strip charts or similar activities) routinely performed for criteria pollutant
    data reported by the agency?  Yes	  No	
    If no, please explain.
(c)  Are procedures for such data audits documented? Yes        No
(d) Are they consistent with the recommendations of Sections 2.3-2.9 of Vol. II of the QA Handbook for Air Pollution
    Measurement Systems?                          •                  '                                 •
    Yes	  No	 If no, why not?.
(e)  What is the frequency and level (as a percentage of data processed) of these audits?

    Poll.            Audit Freq.             Period of Data Audited            % of Data Rechecked
(f)  ldentify_the criteria for acceptable/non-acceptable result from a data provessing audit for each pollutant,  as
    appropriate.

    Pollutant                  Acceptance Criteria                  Data Concentration Level
(g)  Are procedures documented and implemented for corrective actions based on results of data audits which fall
    outside the established limits?  Yes	  No

    If yes, where are such corrective, action procedures documented?
                                                   LF-38

-------
                         Section 2.0.11                64               April 1985



 5.  CORRECTIVE ACTION SYSTEM


 la)  Does the agency have a comprehensive Corrective Action program in place and operational?  Yes
     No
 (b)  Have the procedures been documented? Yes 	  No  	As a part of the agency QA Plan? Yes
     No	As a separate Standard Operating Procedure?  Yes        No	 Briefly describe it or attach a
     copy.                                                              	
 (c)  How is responsibility for implementing corrective actions on the basis of audits, calibration problems, zero/span
     checks, etc. assigned? Briefly discuss.	
 (d)  How does the agency follow up on implemented corrective actions?.
(e) Briefly describe two (2) recent examples of the ways in which the above corrective action system was employed
    to remove a problem area with
     I. Audit Results:
    II.  Data Management:
6. AUDIT RESULT ACCEPTANCE CRITERIA


(a)  Has the agency established and has it documented criteria to define agency-acceptable audit results?
    Yes	  No	


    Please complete the table below with the pollutant, monitor and acceptance criteria.


    Pollutant                     •                   Audit Result Acceptance Criteria
      CO




      03



      NO2
            ^f (continuous}



      SOy






     -A/p
      I TT1T J*
                                                 LF-39

-------
                             April 1985                65                Section 2.0.11


(b)  Were these audit criteria based on, or derived from, the guidance found in Vol. II of the QA Handbook for Air
     Pollution Measurement System, Section 2.0.12?   Yes	/Vo___
     If no, please explain.
     If yes, please explain any changes or assumptions made in the derivation.
     What corrective action may be taken if criteria are exceeded? If possible, indicate two examples of corrective
     actions taken within the period since the previous systems audit which are based directly on the criteria dis-
     s*nccart ah\n\/o
(c)

    cussed above.

    Corrective Action # 1
    Corrective Action #2.
(d)  As a goal, the 95 percent probability limits for precision (all pollutants) and TSP accuracy should be less than
    ± 15 percent. At 95 percent probability limits, the accuracy for all other pollutants should be less than
    ±20 percent. Using a short narrative and a summary table, compare the reporting organization's performance
    against these goals over the last year. Explain any deviations.

NOTE: Precision and accuracy are based on reporting organizations; therefore this question concerns the reporting
organizations that are the responsibility of the agency. A copy of a computer printout has been provided which con-
tains the precision and accuracy data submitted to EMSL for each of the agency's reporting organizations. The print-
out, containing at least the last four completed calendar quarters of precision  and accuracy data, was obtained using
the NADB program NA273. This data should be verified using agency records. If found in error, please initiate correc-
tions. Based on the data provided or corrections thereto, complete the tables below indicating the. number of report-
ing organizations meeting the goal stated above for each pollutant by quarter.

                         {Report, level 2 checks unless otherwise directed by Regional Office.)

    I. Precision Goals

                                                                          Precision
    Pollutant           # of Reporting  Organization          Qtr/Yr      Qtr/Yr     Qtr/Yr     Qtr/Yr


    03    .                         '.'''.



    NO?
    SO,
    CO
    *&*>
    Pb
                                                   LF-40

-------
                         Section 2.0.11                 66                April 1985




     //. Accuracy Goals
                 *


                                                                          Precision
     Pollutant           # of Reporting Organization           Qtr/Yr      Qtr/Yr     Qtr/Yr     Qtr/Yr
    N02




    SO2




    CO


    PM-/0
    Pb
M  To the extent possible, describe problems preventing the meeting of precision and accuracy goals.
                                                  LF-41

-------
                              April 1985
                                                       67
                                             Section 2.0.11
  11.8  Bibliography
    Guideline documents for the
  SLAMS Air Program, arranged in de-
  scending chronological order, the
  most recent ones first.
          Reference
                                              Report Title
  EPA-600/4-83-023
  June 1983
  EPA-600/7-81-010
  May 1981
  EPA-QAMS-005/80
  December 1980

  EPA-600/4-80-030
  June 1980

  EPA-600/4-79-056
  September 1979

  EPA-600/4-79-057
  September 1979

  EPA-600/4-79-019
  March 1979

  EPA-450/4-79-007
  February 1979

  EPA-600/4-78-047
  August 1978
 EPA-450/2-78-037
 July 1978

 EPA-450/3-78-013
 April 1978

 EPA-450/3-77-018
 December 1977

 EPA-600/4-77-027a
 May 1977
 EPA-450/3-77-013
 April 1977

 EPA-450/2-76-029
 December 1976

 EPA-450/2-76-005
 April 1976

^EPA-600/9-76-005
 March 1976

 EPA-450/2-76-001
 February 1976
 Guideline on the Meaning and Use of Preci-
 sion and Accuracy Data Required by 40 CFR
 Part 58 Appendices A and B

 A Procedure for Establishing Traceability of
 Gas Mixtures to Certified  National Bureau of
 Standards SRMs

 Interim Guidelines and Specifications for
 Preparing Quality Assurance  Project Plans

 Validation of Air  Monitoring Data
 Transfer Standards for Calibration of Air Moni-
 toring Analyzers for Ozone

 Technical Assistance Document for the Cali-
 bration of Ambient Ozone Monitors

 Handbook for Analytical Quality Control in
 Water and Wastewater Laboratories

 Guidance for Selecting TSP Episode Monitor-
 ing Methods

 Investigation of Flow Rate Calibration Proce-
 dures Associated with the High Volume
 Method for Determination of Suspended Par-
 ticulates

 Screening Procedures for Ambient Air Quality
 Data

 Site Selection for the Monitoring of Photo-
 chemical Air Pollutants

 Selecting Sites for Monitoring Total Sus-
 pended Particulates

 QA Handbook for Air Pollution Measurement.
 Systems, Vol. II—Ambient Air Specific
 Methods

 Optimum Site Exposure Criteria for SO2 Moni-
 toring

 Aeros Manual Series, Vol. II—Aeros User's
 Manual

 Aeros Manual Series, Vol. V—Aeros Manual of
 Codes

 QA Handbook for Air Pollution Measurement
 Systems, Vol. I—Principles

Aeros Manual  Series, Vol. I—Aeros Overview

-------
                          Section 2.0.11                68                April 1985
	Reference	Report Title	

EPA-450/3-75-077             Selecting Sites for Carbon Monoxide Monitor-
September 1975              ing

APTD-1132                   Quality Control Practices in Processing Air Pol-
March 1973                  lution Samples

47 FR 54912, Dec. 6, 1982;   Amendments to reference methods for SO2,
48 FR 17355, Apr. 22, 1983     TSP and CO in 40 CFR Part 50 Appendices A,
                             B, and C
Proposed amendments to 40 CFR Part 58 are pending.
Proposed revision (Handbook. Vol. II. Sections 2.0.7 and 2.0.9 are pending).

-------
                             June 1984
                                                                        Section 2.O.12
  12.0  AUDIT PROCEDURES FOR USE  BY STATE AND LOCAL AIR MONITORING
                                               AGENCIES
  12.1  Introduction
    Appendix A1 outlines the minimum
  quality assurance requirements for
  state and local air monitoring sta-
  tions (SLAMS). All subsequent revi-
  sions to Appendix A have been in-
  cluded in the preparation of this
  document.2 Quality assurance guide-
  lines for  PSD monitoring are found in
  Appendix B.3
    This section describes performance
  audit procedures for each automated
  and manual monitoring method ref-
  erenced in Appendix A. In additfon,
  quality assurance and quality control
  are defined, standard traceability pro-
  cedures are discussed, and data in-
  terpretation procedures are specified
  relative to the requirements of Ap-
  pendix A.1

  12.2  Quality Assurance and
 Control
   Emphasis''on quality  assurance is
 increasing in the environmental com-
 munity. Since its introduction in the
 manufacturing industry 30 years ago,
 quality assurance has expanded in
 scope to include all phases  of envi-
 ronmental monitoring.
   Quality  assurance consists of two
 distinct and equally important func-
 tions. One function is the assessment
 of the quality of the monitoring data
 by estimating th.eir precision and ac-
 curacy. The  other function is the con-
 trol and improvement of data quality
 by implementing quality control poli-
 cies and procedures and by  taking  -
 corrective actions. These two func-
 tions form a control loop where the
 assessment  indicates when data
 quality is inadequate and where the
.control effort must be increased until
 the data quality is acceptable.
   Each agency should develop and
 implement a quality control  program
 consisting of policies, procedures,
 specifications, standards,.corrective
 measures, and documentation neces-
 sary to:
   1. Provide data of adequate quality
  to meet  monitoring objectives and
  2. Minimize loss of air quality data
  because of malfunctions and out-
  of-control conditions
The selection and degree of specific
control measures and corrective ac-
 tions depend on a number of factors
 such as the monitoring methods and
 equipment, field and laboratory con-
 ditions, monitoring objectives, level
 of data quality required, expertise of
 assigned personnel, cost of control
 procedures, and pollutant concentra-
 tion levels.

 12.3  Standard Traceability
   Traceability is the  process of trans-
 ferring the accuracy  or authority of a
 primary standard to  a field-usable
 standard. Gaseous standards (perme-
 ation tubes and devices and cylin-
 ders of compressed  gas) used to ob-
 tain audit concentrations of CO, S02,
 and NO2 must be working standards
 certified by comparison to NBS-
 SRM's. Traceability protocols are
 available for certifying a working
 standard by direct comparison to an
 NBS-SRM.4-5  Direct use of an NBS-
 SRM is discouraged  because of the
 limited supply and expense. NBS-
 SRM availability and ordering proce-
 dures are given in Reference 6.
   Test concentrations for O3 must be
 obtained by means of an UV photo-
 metric calibration procedure (Subsec-
 tion 12.10.4) or by a certified transfer
 standard.7 Flow measurements must
 be made by an instrument that is
 traceable to an authoritative volume
 or other standard.8-9

 12.4  General Discussion of
 Audit Procedures
   The benefits of a performance
 audit are twofold. From a participant
 standpoint, agencies  are furnished  a
 means of rapid self-evaluation of a


Table 12;1.    Audit Procedures

         Pollutant
       specific monitoring operation. The
       EPA is furnished a continuing index •
       of the validity of the data reported to
       the air quality data bank.
         The performance audit is'used to
       validate and document the accuracy
       of the data generated by a  measure-
       ment system. A list of the specific
       audit procedures which are outlined
       in this section is contained in Table
       12.1. Procedures which use the prin-
       ciples of dynamic dilution,  gas phase
       titration, UV photometry, and flow
       rate measurement are presented.
         The general guidelines for per-
       formance audits are the same for all
       procedures.
         1. A  performance audit  should be
         conducted only if calibration data
         are available for the analyzers or
       •  samplers being audited.
         2. A  performance audit  should be
         conducted only if the site operator
         or representative is present, unless
         written permission is given to the
         auditor before the audit.
         3. Before the audit, a general proce-
         dures protocol, including  the audit
         policy and special instructions
         from  the auditor, should be pro-
         vided to the agency to be audited.
         4. A  signed acknowledgment.that
         the audit has been completed
         should be obtained  from  the sta-
         tion operator.
         5. The auditor should discuss the
         audit results with the site operator
         or representative at the conclusion
         of the audit. A form showing the
       •  audit concentrations, station  re-
         sponses, and other pertinent data
         recorded by the auditor should be
                                           Audit procedure
Sulfur dioxide


Nitrogen dioxide

Carbon monoxide


Ozone

Total suspended paniculate
Dynamic dilution—permeation tube
Dynamic dilution—compressed gas cylinder

Gas phase titration

Dynamic dilution—compressed gas cylinder
Multiple compressed gas cylinders

Ultraviolet photometry

Flow rate measurement

-------
                         Section 2.0.12
                                   June 1984
  given to the site operator or repre-
  sentative; the form must indicate
  that the results are not official until
  the final report is issued. If the site
  operator or representative is not
  on-site at the conclusion of the
  audit, the auditor should contact
  the agency before leaving  the area
  or promptly when returning to the
  base of operations.
  6. The auditor should document the
  verification of his equipment be-
  fore and after the audit; this verifi-
  cation includes calibration and
  traceability data. This information
  and a written record  of the audit
  should be kept in a bound note-
  book in a secure location.
  7. The auditor should  use specific
  procedures that are consistent with
  the performance audit procedures
  manual. Any deviation from these
  must be approved by the agency
  performing the audit.
  8. All audit equipment  and stand-
  ards including standard  gases, per-
  meation tubes, flow measuring ap-
  paratus, and temperature and
  pressure monitors should be refer-
  enced to primary standards.
  9. Verification of the  total audit sys-
  tem output by performing  an audit .
  on calibrated instrumentation
  should be"conducted before the
  audit. The verification instrumenta-
  tion should be calibrated using an
  independent set of equipment and
•  standards.
  10. Upon arrival at the audit site, all
  equipment should be inspected for
  transit damage. Each auditor
  should have a quality control
  checklist or a specified procedure
  that can be used to verify system
  integrity.

 Before  starting the audit,  the auditor
should record the following data: the
site address, operating  agency, type
of analyzer being  audited, zero and
span settings, type of in-station cali-
bration used, and general operating
procedures. These data may  be used
later to determine the cause-of dis-
crepancies between the audit con-
centrations and station  responses.
The auditor should also mark the
data record with a stamp similar to
the  one shown in Figure 12.1 to ver-
ify that the audit was performed and
to prevent the audit data from being
transcribed and mistaken for ambient
monitoring data. Before disconnect-
ing  a monitor or sampler from its
ambient sampling mode, have the
station operator make a note on the
data acquisition system to  indicate
that an audit is being performed.
                           Performance A udit by
                                           PEDCo Environmental, Inc.
                                           11499 Chester Road
                                           Cincinnati, Ohio 45246-0)'00
     Date .

     Start.

     Stop-
                                    Auditor.
Parameter.

Location —
Figure 12.1    Audit identification stamp.
  All station responses should be
converted by the station operator to
engineering units (e.g., ppm or
|j.g/m3) by using the same proce-
dures used to convert the actual am-
bient data. This procedure allows
evaluation of the total monitoring
system—the station operator, equip-
ment, and procedures.
  Upon completion of the audit, all
monitoring equipment must be re-
connected and returned to the con-
figuration recorded before initiating
the audit. Before the auditor leaves
the station, audit calculations should
be performed to ensure that no ex-
traneous or inconsistent differences
exist in the data. Sometimes a
recording mistake is found after leav-
ing the station, and the error cannot
be rectified without returning to the
test site.

12.5   Sulfur Dioxide Audit
Procedure Using Dynamic
Permeation Dilution

12.5.1  Principle—Audit concentra-
tions are generated by a dynamic
system which dilutes an S02 perme-.
ation source with clean, dry air. This
method can be used to audit all com-
mercially available SOa/total  sulfur
analyzers. Several variations in clean,
dry air must be  made to accommo-
date operating characteristics of cer-
tain analyzers. The procedure, its ap-
plicability, precision and accuracy,
and apparatus requirements  are dis-
cussed in the following subsections.

12.5.2  Applicability—The dynamic
dilution  method can be used to sup-
ply SOZ  audit concentrations in the
  range of 0 to 0.5 ppm. Concentra-
 tions for challenging other operating
  ranges such as 0 to 50 ppb, 0 to 0.2
  ppm, 6 to 1.0 ppm, and 0 to 5 ppm
  can also be generated by using this  '
  procedure.

 -12.5.3  Accuracy—The accuracy of
 . the audit procedure should be
  within ±2.5% if the SO2 permeation
  source is referenced and if gas flow
  rates are determined using EPA-
  recbmmended procedures.

  12.5.4  Apparatus—An audit system
  which uses a dynamic permeation di-
  lution device to  generate concentra-
  tions is illustrated in Figure 12.2. The
  eight components of the system are
  discussed below.
    1. Permeation Chamber—A
 .constant-temperature chamber capa-
  ble of maintaining the temperature
  around the permeation tube to an ac-
  curacy of ±0.10°C is required. The
  permeation oven should be equipped
  with a readout that.is sensitive
  enough to verify the temperature of
  the permeation device during normal
  operation.

   2. Flow Controllers—Devices capa-
  ble of maintaining constant flow
  rates to within ±2% are required.
  Suitable flow controllers include
  stainless steel micrometering valves
  in tandem with a precision regulator
  and with mass flow controllers, capil-
  lary restrictors, and porous plug re-
  strictors.
   3. Flowmeters—Flowmeters capa-
  ble of measuring pollutant and dilu-
  ent gas flow rates to within ±2% are .
  required. NBS-traceable soap bubble

-------
                             June 7984
                                    Section 2.0.12
     Clean
      Dry
      Air
                                   Flow
                                 Controller
               Flowmeter
                                  Flow
                                Controller
               Flowmeter
                                                              Output
                                                             Manifold
                           Vent
                              £xtra Outlets Capped
                                When Not in Use
 Figure 12.2.    Schematic diagram of a permeation audit system.
                                                                                  To Inlet of Analyzer
                                                                                    Being Audited
 flowmeters, calibrated mass flow
 controllers or mass flowmeters, and
 calibrated orifice, capillary, and
 porous plug restrictors are suitable.
  4. Mixing Chamber—A glass cham-
 ber is used to mix SO2 with dilution
 air. The inlet and outlet should be of
 sufficient diameter,so that the cham-
 ber is at atmospheric pressure under
 normal operation, and sufficient tur-
 bulence must  be created in the
 chamber to facilitate thorough mix-
 ing. Chamber  volumes in the range
 of 100 to 500 cm3 are sufficient.
 Glass Kjeldahl connecting flasks are
 suitable mixing chambers.
  5. Output Manifold and Sample
 Line—An output manifold used to
 supply the analyzer with an audit at-
 mosphere at ambient pressure
 should be of sufficient diameter to
 ensure a minimum pressure drop at
 the analyzer connection, and the
 manifold miist be vented so that am-
 bient  air will not mix with  the audit
 atmosphere during system opera-
 tions. Recommended manifold mate-
 rials are glass  or Teflon. The sample
 line must be nonreactive and flex-
 ible; therefore. Teflon tubing is pre-
ferred.
 6. Dilution Air Source—The diluent
source must be free of sulfur con-
 taminants and water vapor; clean dry
 air from a compressed gas cylinder
 (Grade 0.1)  may be used. When au-
 diting a flame photometric analyzer,
 a diluent source which contains ap-
 proximately 350 ppm CO2 is re-
 quired. A clean air system may be
 used; however, the system must not
 remove the CO2 from the ambient
 airstream.
  In all cases, the O2 content of the
 diluent air must be 20.9 ±0.2%. Gas
 manufacturers that blend clean dry
 air do not always adhere to the exact
 ambient proportions of 02 and N2; in
 these cases, the O2 content should
 be verified by paramagnetic re-
 sponse.
  7. Sulfur Dioxide Permeation
 Tube—An SO2 permeation tube with
 NBS traceability is used as the pollu- .
 tant source. Permeation rates be-
 tween 0.5 to 1.5  (j.g/min fulfill the au-
 diting requirements. Traceability is
 established  by referencing the per-
 meation device to an NBS-SRM
 (number 1625, 1626, or 1627).

  8.   Permeation Tube Storage—A
storage device capable of keeping
the permeation tube encased in dry
air is required; small cases contain-
ing  Drierite or silica gel will serve
 this purpose. The useful life of a per-
 meation tube will vary among ven-
 dor types {a 9-mo life can be used
 for estimating purposes); low tem-
 perature (2° to 5°C) will prolong the
 tube life. Do not freeze the perme-
 ation tube.

 12.5.5  Procedure—Equipment Setup
 —Remove the permeation tube from
 the storage case, insert it into the
 permeation chamber, and start the
 carrier flow (approximately 50 cm3'
 min) across the  tube. Set the perme-
 ation temperature at  the desired set-
 ting and allow the permeation source
 to equilibrate. For changes of 1° or
 2°C, an  equilibrium period of 3 h
 should suffice. For changes of 10°C
 or when the source is removed from
 low temperature storage, an equi-
 librium  period of 24 h is  advisable.
 Several commercially available per-
 meation calibrators use a carrier flow
 to maintain  a constant temperature
 around the tube during transport. In
 this instance, equilibration is not nec-
 essary because the oven temperature
 is continuously maintained within
 ±0.10°C of the desired permeation
temperature.
  Audit sequence—After all the
equipment has been assembled and

-------
                         Section 2.0.12
                                  June 1984
set up, have the station operator
mark the strip chart recorder to indi-
cate that an audit is beginning. The
auditor's name, start time, date, and
auditing agency should be entered; if
it is not possible to record written
comments on the chart, record the
start and stop times to preclude the
use of audit data as monitoring data.
After recording these data, discon-
nect the analyzer sample  line from
the station manifold and connect it to
the audit manifold, as shown in Fig-
ure 12.3. Cap the sample  port on the
station manifold. (The audit atmos-
phere must be introduced through
any associated filters or sample pre-
treatment apparatus to duplicate the
path taken by an ambient sample.)
Record the analyzer type  and other
identification data  on the  data  form
(Table 12.2J.
  Conduct the audit as shown  in
steps 1-5 below.
  1.  Introduce into the audit mani-
  fold a clean dry air gas at a flow
  rate in excess of 10% to 50% of the
  analyzer sample demand. Allow
  the analyzer to sample  the clean
  dry air until a stable response is
  obtained; that is, until the  re-
  sponse does not vary more than
  ±2% of the measurement range
  over a 5-min period.  Obtain the
  station response and concentration
  from the station operator, and
  record the data in the appropriate
  blanks on the data form.
  2.  Generate SLAMS audit concen-
  trations (which are compatible with
  the analyzer range) as audit  atmos-
  pheres consistent with the require-
  ments in Appendix A.1
                   Station Manifold
                    T  T   T
                     To Analyzers
Audit point
     1
     2
     3
     4
 Concentration range
	(ppm)	
      0.03-0.08
      0.15-0.20
      0.35-0.45
      0.80-0.90
 Generate the concentrations by ad-
 justing the dilution air flow rate
 (FD) and the permeation device air
 flow rate (Fc) to provide the neces-
 say dilution factor. Calculate the
 concentrations as follows.
 [S02] =
         PR X 103
   x 3.82 x 10~4
                     Equation 12-1
 where
   [SOJ = SO2 audit concentration,
           ppm,
      PR = permeation flow rate at
           the specified tempera-
           ture, p.g SO2/min,
      FC = carrier flow rate over the
           permeation tube, stand-
           ard liters/min, and
      FD = diluent air flow rate,
           standard liters/min.

 103 converts liters to  m3, and  the
 3.82 x 10~4 converts  p-g SO2/cm3 to
 ppm SO2 at 25°C and 760 mm Hg.
 3. Generate the highest audit con-
 centration first, and consecutively
 generate audit points of decreasing
 concentration. Allow the analyzer
 to sample the  audit atmosphere
 until a stable response is obtained.
 Obtain the station response and
 concentration from the station op-
 erator, and record the data in the
 appropriate spaces in Table 12.2.
Station
Analyzer


Data
System
(Teletype
\Printout

. in
Volts
                       {Audit Manifold-*-  Exhaust
  4. If desired, additional points at up-
  scale concentrations different from
  those specified in step 2 may be
  generated. Generation of these au-
  dit concentrations plus a  post audit
  clean dry air response will enhance
  the statistical significance of the
  audit data regression  analysis.
  5. After supplying all  audit concen-
  trations and recording all data, re-
  connect the analyzer sample line to
  the station manifold. Make  a nota-
  tion of the audit stop time and
  have the station operator make a
  note on the data recorder to indi-
.  cate the stop time. Have the station
  operator check all equipment to
  ensure that it is in order to  resume
  normal monitoring activities.

12.5.6  Calculations- — Tabulate the
data in Table 12.2 in the appropriate
blank spaces.
  % difference — The % difference is
calculated as follows.
   % Difference =
                                                 x 100,
Flgutt 12.3.    Schematic of configuration utilized in auditing the gas analyzers.
                                             Equation 12-2
                       where
                         CM = the station measured concen-
                              tration, ppm
                         CA = the calculated audit concentra-
                              tion, pprn.

                         Regression analysis — Calculate by
                       the method of least squares the
                       slope, intercept, and correlation coef-
                       ficient of the station analyzer re- •
                       sponse data (y) versus the audit con-
                       centration data (x). These data can
                       be used to interpret the analyzer per-
                       formance.

                       12.5.7  ffeferences— References 4
                       through 6 and 10 and 11 provide ad-
                       ditional information on this S02 audit
                       procedure.

                       12.6   Sulfur Dioxide Audit
                       Procedure Using Dynamic
                       Dilution of a Gas Cylinder

                       12.6.1  Principle — A dynamic dilu-
                       tion system is used to generate SO2
                       concentrations in air for auditing
                       continuous ambient analyzers. The
                       audit procedure consists of diluting a
                       gas cylinder of low S02 concentra-
                       tion with clean dry dilution air. Trace-
                       ability is  established by referencing
                       the gas cylinder to an NBS-SRM.
                       This procedure can be used to audit
                       all commercially available SO2/total
                       sulfur analyzers.
                         Variations in clean dry air must be
                       made to  accommodate operating
                       characteristics of certain analyzers.

-------
                              June 1984
                                                                         Section 2.0.12
 Table 12.2.
 Station 	
Sulfur Dioxide Audit Data Report
 Address
 TA	
 Analyzer
 Calibration standard
 Last calibration date
 Calibration comments
 Zero setting 	
 Span setting  	
 Audit system 	
 Audit standard
 Clean, dry air _
                              mm Hg;
                                                                                             Date
                                                                         Start time
                                                                            mm Hg  Auditor
                                                                   Frequency
                                                                     Serial number
                                                                    .  Span source
                                                                    	 Range
                                                Data acquisition system
                                               	  Recorder
                                Bubble flowmeter serial number
                        ;  P
                                             psig;  [  ] =.
                                                                           Catalytic oxidizer     Yes
                                                                                         ppm
                                                                                         No
 Flow correction:

 Dilution air response
     Other response  _
  	 P   \
A   "W^OX
760 mm I
 Audit PointT
     Dilution flow measurement
     Volume	
        T, 	.	
        T2 	
        T3	
    Analyzer response _
        Other response
                       298 K
                      TA + 273 ,
                            % Chan;
                                         mm
                             % Chart;
                                                          Flowmeter
                                                               Volume
                                           Audit concentration
                                           	  VDC;  	
 cm-3
 min
 ppm
 ppm
Audit'Point II
    Dilution flow measurement
    Volume 	
        T,  	
   Analyzer response _
       Other response
                      cm-
                                                       mm
                            % Chart;
                                                         Flowmeter
                                           Audit concentration
                                          	 VDC;  	
 min
ppm
ppm

-------
                         Section 2.0.12
                                                                     June 1984
Table 12.2  (continued)
Audit Point III
    Dilution flow measurement
    Volume	
        7, 	
        7>	
        r3	:—
    Analyzer response
        Other response
Audit Point IV
    Dilution flow measurement
    Volume	'.
       T, 	
    Analyzer response —
        Other_response .

Audit Point V
   Dilution flow measurement
   Volume	
       T, 	
       Ta	
    Analyzer response
       Other response
MethnH
                                                                     Flowmeter
                                                     mm
                                                                          Volume
       (*X*f
                                                               Audit concentration
                                          % Chart;
                                    rm
                                                                     Flowmeter
                                                     mm
                                                                          Volume
                                                                    (*)(**
                                                               Audit concentration
                                         % Chart;
                                    Cm
                                                     mm
                                                                    (CF){^
                                         % Chan;
          ' Volume
            , T
 Audit concentration
	  l/oc;  	
                                                                                                   cmj
                                                                                                   min
                                                                                                  ppm
                                                                                                   cmj
                                                                                                   min
                                                                                                  ppm
                                                                                                  cnv*
                                                                                                  min
                                                                                                 ppm
Permeation temperature
                               °r-
                                                                     Permeation rate
                                                                                                ^g/min

-------
                              June 1984
                                                                          Section 2.0.12
 Table  12.2  (continued)
               Gas flow rates,
                std cm3/min
  Point
 number
QC
QD
    Audit
concentration,
    ppm
                                          Analyzer response
                                                                                            Difference
                                                        ppm
    mV/
%  chart
Analyzer-audit,
     ppm
 Regression analysis [audit concentration (x) vs. analyzer response (y)J
                  y = mx + b

             Slope (m) 	

           Intercept (b) 	

         Correlation (r) 	

 Comments:
The procedure, its applicability, accu-
racy, and apparatus requirements are
discussed in the following subsec-
tions.

12.6.2   Applicability—Dynamic dilu-
tion can be used to supply SOj audit
concentrations in the range of 0 to
0.5 ppm. Concentrations for challeng-
ing other operating ranges such as 0
to 50 ppb, 0 to 0.2 ppm, 0 to 1.0
ppm, and 0 to 5 ppm can also be
readily generated by using this pro-
cedure.

12.6.3   Accuracy—The accuracy of
the audit procedure should be within
±2.5% if the SO2 gas cylinder con-
centration is referenced  and if gas
flow rates are  determined  using EPA-
recommended procedures.

12.6.4   Apparatus—An audit system
which uses a dynamic dilution device
to generate audit concentrations is il-
lustrated in Figure 12.4. The seven
                       Clean
                        Dry
                        Air
Flow
Controller


Flowmeter

Fo f 	 -^

Mixing
Chamber
                        Std

                       (50 ppm)
Flow
Controller


Flowmeter


                                                                    FP
                              Vent •
                              Extra Outlets Capped
                                When Not in Use
Output
Manifold
a
^
Ft
'I'
                                                                    To Inlet of Analyzer
                                                                      Being Audited
                                     Flgur* 12.4    Schematic diagram of a dilution audit system.

-------
                         Section 2.0.12
                                                                       June 1984
components of the device are dis-
cussed below.
  1. Gas Cylinder Regulator—A
stainless steel gas regulator is ac-
ceptable. A low deadspace, two-
stage regulator should be used to
achieve rapid equilibration. A purge
assembly is helpful.
  2. Flow Controllers—Devices capa-
ble of maintaining constant flow
rates to within ±2% are required.
Suitable flow controllers include
stainless steel micrometering valves
in tandem with a precision regulator,
mass flow controllers, capillary re-
strictors, and porous plug restrictors.
  3. Flowmeters—Flowmeters capa-
ble of measuring pollutant and dilu-
ent gas flow rates to within ±2% are
required. NBS-traceable soap bubble
flowmeters, calibrated mass flow
controllers or mass flowmeters, and
calibrated orifice, capillary, and
porous plug restrictors are suitable
for  flow determination.
  4. Mixing Chamber—A glass or
Teflon chamber is used to mix the
S02 with dilution air. The inlet and
outlet should be of sufficient diame-
ter so that the chamber, is at atmos-
pheric pressure under normal opera-
tion, and sufficient turbulence must
be created in  the chamber to facili-  .
tate thorough mixing. Chamber vol-
umes in the range of 100 to 500 cm3
are sufficient. Glass Kjeldahl connect-
ing flasks are suitable mixing cham-
bers.
  5. Output Manifold and Sample
Line—An output manifold used to
supply the analyzer with an audit at-
mosphere at ambient pressure
should be of sufficient diameter to
ensure a minimum pressure drop at
the analyzer connection, and the
manifold must be'vented so that am-
bient air will not mix with the audit
atmosphere during system opera-.
tions. Recommended manifold mate-
rials are glass or Teflon. The sample'
line must be nonreactive and flex-
ible; therefore, Teflon tubing is pre-
ferred.
  6. Dilution Air Source—The dilu-
ent source must be free of sulfur
contaminants and water vapor; clean
dry air from a compressed  gas cylin-
der (Grade 0.1) may be used. When
auditing a flame photometric ana-
lyzer, a diluent source which contains
approximately 350 ppm C02 is re-
quired. A clean air system may be
used; however, the system must not
remove the C02 from the ambient
airstream.
  In all cases, the 02 content of the
diluent source must be 20.9 =0.2%.
Gas manufacturers that blend the
clean dry air do not always adhere to
the exact ambient proportions of O2
and N2; in these cases, the 02 con-
tent should be verified by paramag-
netic response.
  7. Sulfur Dioxide Gas Cylinder—A
compressed gas cylinder containing
50 to 100 ppm S02 in air is used as
the dilution source. This cylinder
must be traceable to an NBS-SRM
(number 1661, 1662, 1663, or 1664).

12.6.5  Procedure—Equipment
setup—Assemble the audit equip-
ment as required, and verify that all
equipment is operational. If a dilution
air system equipped with  a catalytic
oxidizer is used, allow the oxidizer to
warm up for 30 min. Connect the gas
regulator to the SO2 cylinder, and
evacuate the regulator as  follows.
  1. With the cylinder valve  closed,
  connect a vacuum pump to the
  evacuation outlet on the regulator
  and start'the pump.
  2. Open and close the evacuation
  port.
  3. Open .and close the cylinder
  valve.
  4. Open and close the evacuation
  port.
  5. Repeat steps 2 through 4 five
  more times to be sure all 02 impu-
  rities are removed from the regula-
  tor.
If the regulator does not have an
evacuation port but has a supported
diaphragm, the procedure can be
conducted at the gas exit  port.
  For regulators that do not have an
evacuation port but have an unsup-
                  Station Manifold
ported diaphragm, use the following
procedure:
  1.  Connect the regulator to the-
  cylinder, and close the gas exit
  port.
  2.  Open and close the cylinder valve
  to  pressurize the regulator.
  3.  Open the gas exit port, and allow
  the gas to purge the regulator.
  4.  Repeat steps  2 and 3 five more
  times; then close the gas exit port,
  and open the cylinder valve.  (The
  regulator should remain  under
  pressure.) Connect the gas cylinder
  to  the audit device.
Repeat the procedure for each  cylin-
der.
  Audit sequence—Before discon-
necting the analyzer from the station
manifold, mark the data recorder to
indicate that an audit is beginning.
The auditor's name, start time, date,
and auditing organization should be
recorded. If it is not possible to
record written comments, the start
and stop times should be recorded to
preclude the use of audit data as
monitoring data. After recording
these data, disconnect the  analyzer
sample line from the station mani-
fold, and connect it to the audit man-
ifold, as shown in Figure 12.5. Cap
the sample port on the station  mani-
fold. (The audit atmosphere must be.
introduced through any associated
filters or sample  pretreatment ap-
paratus to duplicate the path taken
by an ambient sample.) Record the
analyzer type and other identification
data on the data  form  (Table 12.3).
  Conduct the audit by following
steps 1 through 5 below.
                    T   T   T   T
                    To Analyzers
»
i m
Station
Analyzer
n


Data
Acquisition
System

                       (Audit Manifold-*- Exhaust
Figure 12.5   Schematic of configuration utilized in auditing the gas analyzers.

-------
                            June 1984
                                                                      Section 2.0.12
 Table 12.3.
 Station
              SO2 Audit Data Report
                                                                                          Date
 Address
                                                                                     Start time
                                           mm Hg;
                                                                         mm Hg Auditor
Analyzer
 Calibration standard
 Last calibration date
                                                                 Frequency
                                                                                 Serial number
                                                                                        source
                                                                                        Range
 Calibration comments
 Zero setting 	
 Span setting 	
 Audit system 	
Audit standard
Clean, dry air _
Flow correction:
    Dilution air flow
    Volume 	
       r,	
       TZ~	
       7-5	
                                                             Data acquisition system
                                                            	  Recorder
                                             Bubble flowmeter serial number
                      PA - PH2o\   (  298 K
                       760mm I
273
                                     cm
                                                       mm
                                                                  psig;  I  1 =.
                                                                         Catalytic oxidizer     Yes
                                                                                                      No
                                                                         VV
                                                                                                         3-
                                                               cm
                                                               min
Dilution air response
    Other response  _
Audit Point I
    Pollutant flow measurement
    Volume 	
    Analyzer response _
       Other response
                                          % Chart;
                                     cmj
                                                      mm
                                                                          'DO
                                          % Chart;
                                                                       Flowmeter
                                                                       fc\(Volume\ =
                                                                       \ F/\  f   I
                         Audit concentration
                        	 VDC;  	
 cmj
 min
ppm
ppm

-------
                         Section 2.0.12
                                                    10
                                                                    June 1984
 Audit Point II
     Pollutant flow measurement
     Volume	
        T2
        T3
    Analyzer response
        Other response
 Audit Point III
    Pollutant flow measurement
    Volume
Audit Point IV
    Pollutant flow measurement
    Volume	.	
       r,	
       T2	
    Analyzer response _
       Other response
Audit Point V
    Pollutant flow measurement
    Volume	
       r,	
   Analyzer response _
       Other response
                                    cm-3
       Flowmeter
                                                     mm
                                                                     (r \(v°lume\ -
                                                               Audit concentration
                                    Cm
                                                     mm
                                                                    /c\/Volume\ =
                                         % Chart;
 Audit concentration
	 VDC;. 	
                                                                    Flowmeter
                                                    mm
                                                                    I r \(Volume\ _
                                                                    (CF)(~J~)--
                                         % Chan;
 Audit concentration
	  VDC;  	
                                     C/T7J
                                     min
                                    ppm
                                    ppm
TI
T,

Analyzer response
•tf -
Other response


(c,\(Volume\
\ r)\ T 1
Audit concentration
v.rh*rf ' i/ .

cm3
mm
nnm

	 : 	 HfJI11
 cmj
 min
ppm
ppm
 cmj
 min
ppm
ppm

-------
                             June 1984
                                                       11
                                                                         Section 2.O.12
 Regression analysis [audit concentration (x) vs. analyzer response (y)]

            Slope (m) 	

          Intercept (b) 	

        Correlation (r) 	

 Comments:
Auditor
Audit method
Zero setting
 Station calibration source
                   Span setting
                                     Equivalency reference no.
Point
number
                     Flow rates
                                                                 Analyzer response
Pollutant
cm3/mm
  Total,
cm3/mm
    Audit
concentration,
    ppm
                                                                ppm
% Chart
 orMV
                                                                               Difference
Analyzer
 audit,
 •ppm
Regression analysis [audit concentration (x) vs. analyzer response (y)]

                 y = mx + b

            Slope (m) 	

         Intercept (b) 	

        Correlation (r) 	

Comments:

-------
                        Section 2.0.12
                  12
                                                                      June 1984
 1. Introduce into the audit mani-
 fold a clean dry air-gas at a flow
 rate in excess of 10% to 50% of the
 analyzer sample demand. Allow
 the analyzer to sample the clean
 dry air until a stable response is
 obtained; that is, until the re-
 sponse does not vary more than
 ±2% of the measurement range
 over a 5-min period. Obtain the
 station response and concentration
 from the station operator and
 record the data in the appropriate
 blanks on the data form.
 2. Generate the SLAMS audit con-
 centrations (which are compatible
 with the analyzer range) as audit
 atmospheres consistent with the
 requirements in Appendix A.1
Audit point
1
2
3
4
Concentration range
(ppm)
0.03-0.08
0.15-0.20
0.35-0.45
0.80-0.90
 Generate the audit concentrations
 by adjusting the pollutant flow rate
 (FPJ and the total flow rate (FT) to
 provide the necessary dilution fac-
 tor.
   Calculate the audit concentration
 as follows.


 (S02J » p£ x [S02)STD Equation 12-3
 where
     (SO2J  » audit concentration of
            SO2, ppm,
        Fp = pollutant  flow rate,
            cm3/min
        FT « total  flow rate, cm3/
            min [equal to the sum
            of the pollutant flow
            rate (FP) and the dilu-
            tion flow rate (FD)],
            and
   (SOjlsro ^concentration of the
            standard cylinder,
            ppm.

 3. Generate the highest audit con-
 centration first, and consecutively
 generate audit points of decreasing
 concentration. Allow the analyzer
 to sample the audit atmosphere
 until a stable response is obtained.
 Obtain the  station response and
 concentration from the station op-
 erator, and record the data in the
 appropriate spaces in Table 12.3.
4. If desired, additional points at up-
scale concentrations different from
those specified in step 2 may be
generated.  Generation of these au-
dit concentrations plus a post audit
   clean dry air response will enhance
   the statistical significance of the
   audit data regression analysis.
   5.  After supplying all audit sample
   concentrations and recording all
   data, reconnect the analyzer sam-
   ple line to the station manifold.
   Make a notation of the audit stop
   time. Have the station operator
   make a note on the data recorder
   to  indicate the  stop time, and
   check all equipment to ensure that
   it is in order to resume normal
   monitoring activities.

 12.6.6  Calculations — Record the
 data  in Table 12.3 in the appropriate
 spaces.
   % difference — The % difference is
 calculated as follows.
    % difference =
                                                        _ s*
                                                             x 100,
                      Equation 12-4
 where
   Cto = the station-measured concen-
        tration, ppm, and
   CA = the calculated audit concentra-
        tion, ppm

   Regression analysis — Calculate by
 the method of least squares the
 slope, intercept, and correlation coef-
 ficient of the station analyzer re-
 sponse data (y) versus the audit con-
 centration data (x). These data can
 be used to interpret the analyzer per-
 formance.

 12.6.7  References — References 4
 through 6 and 10 and 11 provide ad-
 ditional information on this SO2 audit
 procedure.


 12.7   Nitrogen Dioxide Audit
 Procedure Using  Gas  Phase
 Titration

 12.7.1   Principle — The  auditing pro-
 cedure is based on the  gas phase re-
 action between NO and O3

 NO + O3 -» NO2 + O2.  Equation 12-5

The generated NO2 concentration is
 equal to the NO concentration con-
sumed by the reaction of O3 with ex-
cess  NO.
  The NO and NOX channels of the
 chemiluminescence  NOX analyzer are
 audited with known  NO concentra-
 tions produced  by a dynamic dilution
 system  which uses clean dry air to
 dilute a gas cylinder containing NO
 in nitrogen. After completion of the
 NO-NOX audits, stoichiometric mix-
tures of NO2 in combination with NO
are generated by adding 03 to
 known NO concentrations. These
 audit data are used to evaluate the
 calibration of the NO-NOX-NO2 ana-
 lyzer channels and to calculate ana-
 lyzer converter efficiency.

 12.7.2 Applicability—The procedure
 can be used to supply audit concen-
 trations of NO-N02-NOX in  the
 range of 0.010 to 2.0 ppm.

 12.7.3 Accuracy—The accuracy of
 the audit  procedure should be within
 ±2.5% if the  NO gas cylinder concen-
 tration is  referenced and  if the gas
 flow rates are determined by using
 EPA-recommended procedures.

 12.7.4 Apparatus—Audit system—A
 typical gas phase titration system is
 illustrated in  Figure 12.6.  All connec-
 tions and components downstream
 from the O3 generator and the pollu-
 tant source must be constructed of
 nonreactive (glass or Teflon) mate-
 rial. The seven components  of the
 system are discussed below.
  1. Flow Controllers—Devices capa-
 ble of maintaining constant flow
 rates to within ±2% are required.
 Suitable flow controllers include
 brass (for air) or stainless steel (for
 NOX) micrometering valves in
 tandem with  a precision regulator,
 mass flow controllers, capillary re-
 strictors, and porous plug restrictors.
  2. Flowmeters—Flowmeters capa-
 ble of measuring pollutant and dilu-
 ent gas flow rates to within ±2% are
 required.  NBS-traceable soap bubble
 flowmeters, calibrated mass flow
 controllers or mass flowmeters, and
 calibrated orifice, capillary, and
 porous plug restrictors are all suit-
 able for flow  determination.
  3. Gas Cylinder Regulator—A non-
 corrosive  two-stage stainless steel -
 regulator with an evacuation port is
 suggested.
  4. Ozone Generator—An O3 gener-
 ator that produces a stable concen-
 tration is required during  the gas
 phase titration sequence of the audit.
 An  ultraviolet lamp generator is rec-
 ommended.
  5. Reaction Chamber—A glass
 chamber used for the quantitative re- •
 action of O3 with  NO should have
 sufficient volume, 100 to 500 cm3, for
 the residence time to be s2  min.
 Elongated glass bulbs such as Kjel-
 dahl connecting flasks are suitable.
  6.  Mixing Chamber—A glass or
Teflon chamber is used to mix the
NO,  NO2, or O3 with dilution air. The
inlet and outlet should be of suffi-
cient diameter so that the chamber is
at atmospheric pressure under nor-
mal operation, and sufficient turbu-.

-------
                              June 1984
                 13
                                                                        Section 2.0.12
o
Cleat
Dry
Air
kWWH
n
\
}
N
St
— *•
^•^H
a
4—1
•N
0
d
Flow
Controller
- Flowmetar

Flow
Controller
51==
- Flowmeter

1 Flow
Controller
Vent + 	
Extra Outlet
When Not
	 »

Fo
Roaction
\Fxo


Output
Manifold
1"
Mixing
Chamber
Ft
ii U, I," -1
s Capped )f
'" Us* To Inlet of Analyzer
Being Audited
  Figure 12.6   Schematic diagram of a gas phase titration audit system.
  lence must be created in the cham-
  ber to facilitate thorough mixing.
  Chamber volumes in the range of
  150 to 250 cm3 are sufficient. Glass
  Kjeldahl  connecting flasks are suit-
  able mixing chambers.
   7.  Output Manifold and Sample
  Line—An output manifold used to
  supply the analyzer with an audit at-
  mosphere at ambient-pressure
  should be of sufficient diameter to
  ensure a minimum pressure drop at
  the analyzer connection, and the
  manifold must be vented so that am-
  bient air  will not mix with the audit
  atmosphere during system opera-
  tions. Recommended manifold mate-
  rials  are glass or Teflon. The sample
  line must be nonreactive and flex-
  ible;  therefore. Teflon is preferred.
   Dilution air system—Clean dry air
  from a compressed gas cylinder
  (Grade 0.1) is a suitable source for
  dilution air; however, if large vol-
  umes of clean dry air (s>5 liters/min)
  are required, purified compressed air
  is preferred. The clean dry air must
  be free of contaminants such as NO,
  NO2, 03 or reactive hydrocarbons
 that would cause detectable re-
 sponses on the  NOX analyzer or that
 might react with NO or NO2 in the
 audit system. The air can be purified
 to meet these specifications  by pass-
 ing it through silica gel for drying, by
 treating it with  03 to convert any NO
.to N02, and by passing it through ac-
hivated charcoal (6-14 mesh) and a
 molecular sieve (6-16 mesh, type 4A)
 to remove NO2, 03, or hydrocarbons.
   Silica gel maintains its drying effi-
 ciency until it has absorbed 20% of
 its weight; it can be regenerated in-
 definitely at 120°C. Addition of cobalt
 chloride to the surface of the gel pro-
 vides a water absorption indicator. A
 transparent drying column is recom-
 mended. The activated charcoal and
 molecular sieve  have a finite absorp-
 tion capability; because  it is difficult
 to determine when the capability has
 been exceeded,  both should be re-
 placed either before each audit or.
 after 8 hrs of use.
  Nitric oxide gas cylinder—A com-
 pressed gas cylinder containing 50 to
 100  ppm NO in N2 is used as the NO
 dilution source. This cylinder must
 be traceable to an NBS-SRM (num-
 ber 1683,  1684, 1685, 1686, or 1687).
12.7.5  Procedure—Equipment
setup—Assemble the audit equip-
ment as required, and verify that all
equipment is operational. If a clean,
dry air system  equipped with a cata-
lytic oxidizer and/or 03 lamp is used,
allow the oxidizer and/or 03 lamp to
warm up for 30 minutes. Connect the
gas regulator to the NO cylinder, and
evacuate the regulator.as follows:
  1.  With the  cylinder  valve  closed,
  connect a vacuum pump to the
  evacuation outlet on the regulator,
  and start the pump.
  2.  Open and close the evacuation
  port.
  3.  Open and  close  the  cylinder
  valve.
   4. Open and  close the evacuation
   port.
   5. Repeat steps 2 through 4 five
   more times to be sure all O2 impu-
   rities are removed from the regula-
   tor.
 If the regulator does not have an
 evacuation port but has a  supported
 diaphragm, the procedure can be
 conducted at the gas exit port.
   For regulators that do not have an
 evacuation port but have an unsup-
 ported diaphragm, use the following
 procedure:
   1. Connect the regulator  to the
   cylinder, and close the gas exit
   port.
   2. Open and close the cylinder valve
   to pressurize the regulator.
   3.  Open the gas exit port, and"
   allow the gas to  purge the regula-
   tor.
   4. Repeat steps  2  and 3 five more
   times, close the gas exit port, and
   open the cylinder valve.  Connect
   the dilution air source and NO
   cylinder to the audit device as
   shown in Figure  12.6. Use  1/8-in.
   o.d. tubing of minimum  length for
   the connection between  the NO
   cylinder and the audit device.
   Dynamic parameter specifica-
tions — The flow conditions used in
the GPT audit system are selected to
assure a complete N0-03  reaction.
The gas flow rates  must be adjusted
according to the following  relation-
ships:
    PR = [NOJnc x tR s> 2.75 ppm-min,
                      Equation 12-6
(NO]RC
                    rNO
FO
                       FNO
                      Equation 12-7
           "HC
                   2 min,
  ,   R   FO + FNO
                      Equation 12-8
where
       PR=dynamic parameter
           specification, determined
           empirically, to ensure
           complete reaction of the
           available O3/ ppm-min,
   [NO]RC = NO concentration in the
           reaction chamber, ppm,
       tR = residence time of the  re-
           actant gases in the reac-
           tion chamber,  min,
         i = concentration of the NO
           gas cylinder, ppm,
     FNO = NO flow rate, standard
           cm3/min,
       FO = 03 generator air flow rate,
           standard cm3/min, and
     VRC= volume of the reaction
           chamber, cm3.

-------
                          Section 2.0.12
                                                      14
                                                                        June 1984
   The flow conditions to be used in
 the GPT audit system are selected
 according to the following sequence:
   1. Determine FT, the total flow rate
   required at the output manifold (FT
   * analyzer(s) demand plus 10% to
   50% excess).
   2. Determine FNO, the flow rate of
   NO required to generate the lowest
   NO concentration required at the
   output manifold during the GPT
   (approximately 0.15 ppm).
0.15 x FT
 (NOJsro
                               12-9
   3. Measure the  system's  reaction
   chamber volume; must be  in the
   range of approximately 100 to 500
   cm3.
   4. Compute Fo.
   Fo
                    FNO x VRC
                 2.75
                     Equation 12-10

   5. Compute tR, using Equation 12-8;
   verify that tR s 2 min.
   6. Compute F0.

   FO » FT - FO ~ FNO  Equation 12-11
   where
     FO ™ diluent air flow, standard
       cm3/min.

   Adjust FQ to the value determined
 above. FO should not be further ad-
 justed during the NO-NOX or NO2
 audit procedures; only FNO (or FD)
 and the 03 generator settings are ad-
 justed during the course of the audit.

   Audit sequence—After all the
 equipment has been assembled and
 set up, have the station operator
 mark the strip chart recorder to indi-
 cate that the audit is beginning. In-
 formation such as the auditors'
 name, start time, date, and auditing
 organization should be entered. If it
 is not possible to enter written com-
 ments, the start and stop times
 should be recorded  to precluoe the
 use of audit data as monitoring data.
 After recording  the data, disconnect
 the analyzer sample line from the
 station manifold, and connect it to
 the audit manifold, as shown in Fig-
 ure 12.7. Cap the sample port on the
 station manifold. The audit atmo-
 sphere must be introduced through
 any associated filters or sample pre-
treatment apparatus to duplicate the
path taken by an ambient sample.
Record the analyzer  type and other
identification data on the data form
(Table 12.4).
                                           Station Manifold    A

                                             T   T  T   T
                                              To Analyzers
                              Test
                           Atmosphere






Station
Analyzer




Data
System

                                                           -{.A udit Manifold -*• Exhaust
                                      Figure 12.7    Schematic of configuration utilized in auditing the gas analyzers.
                            Conduct the NO-NOX and NO2
                          audits as follows:
                            NO-NOX Audit— The NO-NOX
                          audit involves generating concentra-
                          tions to challenge the calibration of
                          the NO and NOX channels of the ana-
                          lyzer. Data collected during this audit
                          are used to construct a calibration
                          curve that will be used later for cal-
                          culating the NO2 audit concentra-
                          tions.
                            NO-NOX Audit Procedure —
                          1. Introduce clean dry air into the
                          audit manifold at a flow rate in ex-
                          cess of 10% to 50% of the analyzer
                          sampje demand. Allow the analyzer
                          to sample the clean dry air until a
                          stable response is obtained; that is,
                          until the response does not vary
                          more than  ±2% of the measurement
                          range over a 5-min period. Record
                         the readings for the NO, NOX, and
                          N02 channels, and have the station
                         operator report the audit responses
                         in concentration units. Record these
                         data and the responses of all three
                         channels  in table 12.4.
                           2. Generate upscale NO audit con-
                         centrations corresponding to 10%,
                         20%, 40%, 60%, and 90% of the full-
                         scale range of the analyzer by adjust-
                         ing the flow rate of the NO standard.
                         For each audit concentration level
                         generated, calculate the NO concen-
                         tration.
       FP
(NO) =  ; x
                                             Equation 12-12
                         where
                             [NO] = NO-NOX audit concentra-
                                    tion, ppm (the NO2 im-
                                    purity in the stock stand-
                                    ard  should be
                                    negligible),
      Fp = pollutant flow rate, cm3/
          min,
      FT = total flow rate, cm3/min,
          and
        = concentration of the
          standard cylinder, ppm.

     NOTE:  Alternatively, the upscale
     NO audit concentrations may be
     generated by maintaining a con-
     stant  pollutant, flow rate (FP) and
     varying the dilution air flow rate
     (FD). In this case, the entries for
     dilution air flow and pollutant
     flow in Table 12.4 should be re-
     versed and clearly  indicated.
  3. Generate the lowest audit con-
centration level first and consecu-
tively generate audit points of in-
creasing concentration. Allow the
analyzer to sample the  audit atmo-
sphere until a stable response is ob-
tained. Record the audit concentra-
tion.
  Obtain the station response and
concentration from the  station opera-
tor for the NO, NOX, and N02 chan-
nels, and  record the data  in the  ap-
propriate  spaces in Table 12.4.
  4. Prepare audit calibration curves
for the NO and NOX channels by
using least squares. Include the  zero
air  points. (The audit concentration is
the x variable; the analyzer response
in % chart is the y variable.) The NO
audit calibration curve will be used
to determine the actual  audit concen-
trations during the generation of the
NO2 atmospheres.
  The NOX audit calibration curve
will be used to determine NO2 con-
verter efficiency.
  NO2 Audit—The N02  audit in-
volves generating N02 concentra-

-------
                            June 1984
                                                     15
                                                                       Section 2.O.I2
 Table 12.4.
 *faf/on 	
             Gas Phase Titration Audit Data Report
                                                                                         Date
Address
TA	
                                                                                    Start time
                                         mm Hg;
                                                   mm Hg;  Auditor
Analyzer
 Calibration standard
 Last calibration date
                                                                Frequency
                                                            Serial number
                                                           .  Span source
                                                           	 Range
 Calibration comments
 Flow settings
Span settings
 Zero settings
Other settings _
Audit system 	
                              NO
                              NO
                                  NOX
                                  NOX
NO2
N02
Audit standard
                                       	  Bubble flowmeter serial number	
                                       P	 psig;  '[  ] =.
Clean, dry air supply
Flow correction:
Dilution air flow:
   Volume 	
      T, 	
PA ~ PH2o
 760mm
                                   298 K
                                  TA + 273
                                                                       Flowmeter
                                                       mm
                                                                            Volume
                                                                                                      ppm
                     (C)
                                                                                                      cmj
                                                                                                      min
Ozone generator flow:
    Volume 	
        7,  	
        T2	.	
        T3  	
                                    cm0
                                                                      Flowmeter
                                                     mm
                                                                           Volume
                                                                                                     cm-3
                                                                                                     min
Analyzer response clean dry air
                                                    % Chan
                                                                     'DC
                                              NO
                                             NOX
                                             NO2
                                                                                                  ppm

-------
                        Section 2.0.12
                                                  16
                                                                  June 1984
Table 12.4  (continued)
PART I.  NO-NOX AUDIT
NO-NOX Audit Point I (10%)
    Pollutant flow measurement
    Volume  _
        T2
        T3
                                   cnr3
                 mm
                                Flowmeter
                                     yolume
                                                            NO, NOX audit concentration .
                                     cnr1
                                     min
                                    ppm
 Analyzer response
               % Chart
      VDC
                                            NO
                                           NOX
                                           NO2
                                                          ppm
NO-NOX Audit Point II (20%)
    Pollutant flow measurement
    Volume _
        T2
        T3
Analyzer response
cm-2
        Flowmeter
                 mm
                                     Volume
                                     -
              % Chart
NO, NOX audit concentration .
     VDC           (   )
                                           NO
                                           NOX
                                           NO,
cm-3
min
ppm
                                                                                            ppm
NO-NOX Audit Point III (40%)
         *
    Pollutant flow measurement
    Volume	
       r,		
       T2	
       T3	
Analyzer response
cm0
                                Flowmeter
                mm
              % Chart
NO, NOX audit concentration
     VDC          (    )
                                           NO
                                          NOX
                                          NO2
                                                             cmj
                                                             min
                                                                                               ppm
                                                                                            PPm

-------
                           June 1984
17
                        Section 2.0.12
Table 12.4  (continued)
NO-NOX Audit Point IV (60%)
    Pollutant flow measurement
    Volume _
       T2
Analyzer response
                                             NO
                                            NOX
                                            NO2
  mm
% Chart
                   Flowmeter
                                                                          / Volume'
                 NO, NOX audit concentration .
                       VDC         .  (    )'
                                                                                                    cm-3
                                                                                                    min
                                                                                                    ppm
                                                                                                ppm
NO-NOX Audit Point V (90%)
    Pollutant flow measurement
    Volume 	
Analyzer response
                                             NO
                                            NOX
                                            NO2
      NO-NOX audit calibration equation (y = mx + b)
NO audit concentration (x)
  vs. analyzer response in
  % chart (y)
                   Slope (m) =.
                 Intercept (b) .=.
                Correlation- (r) =.
                                                     mm
% Chart
                   Flowmeter
                                                                        \ / Volume-
                            A   T
                 NO, NOX audit.concentration _^.
                       VDC           (    )
    NOX audit concentration (x)
      vs. analyzer response in
      % chart (y)
                        Slope (m) =.
                      Intercept (b) =.
                    Correlation (rj —.
                                                                                                    min
                                                                                                    ppm-
                                             ppm
PART II.  NO2 AUDIT
NO2 Audit Point I
 Analyzer response
% Chart
                                      NO
                                      NOX
          VDC
                                      O3   generator setting =
                                             I    1*     ORIG
                                            	  ppm
                                            	  ppm

-------
                          Section 2.0.12
                                                     18
                         June 1984
 Table 12.4  (continued)
                                               % Chart
                                        NO
                                        NOX
                  VDC
                                                                  = [NO]»ORIC - (NO]*REM = .
REM
ppm
ppm
ppm
                                               % Chart
                  VDC
                                                                                            ppm
                                        NO,
 N02 Audit Point II
 Analyzer response
% Chan
                                                               VDC
                                       NO
                                       NO*
                                            I    ]*     ORIG
                                            	'ppm
                                            	  ppm
                                         O3 generator setting =.
  "Calculated concentration from NO or NOX audit calibration equation (y = mx + b).
 N02 Audit Point III
 Analyzer response
% Chart VDC ( ' ) 11*
A/n
NOX
[NO,!* = //VO/*CfflG - fNO]*fjEM =
% Chart VDC ( ) ppm
NO,
% Chart VDC < ) [' ]*
NO
NO..
O7 aenerator setting =
% Chart VDC ( ) [ ]*
NO
NOX
[N0,l* = fNOl*ORlr; - [NOI*nEM =
% Chart VDC ( ) 11*
NO,
REM

nnm
DOm


ORIC



REM


oom


'Calculated concentration from NO or NOX audit calibration equation (y = mx + b).

-------
                         June 1984
       19
Section 2.0.12
Table 12.4  (continued)
NO2 Audit Point IV
Analyzer response
                                             % Chart
                   VDC
                                      NO
                                      NOX
                                        O3 generator setting
                       [   1*     ORIG
                                             % Chan
                                      NO
                                      NOX
                   VDC           (   >         [   1*     REM
               	    	    	  ppm
               	.	    	:—  ppm
               IN02]A = [NO]*ORIC - (NO]*REM =	  ppm
                                             % Chart.
                                      NO2
                                                              /DC '
NO2 Audit Point V
Analyzer response
                                             % Chan
                                      NO
                                      NOX
                   V,
                                                              DC
                                       O3 generator setting = .
                       [   ]*    ORIG
                      	• ppm
                      	  ppm
                                      NO
                                      NOX
  % Chart          Voc           (    )          [   }*    nEM
		   	;	   	  ppm
	    	   	   	  ppm
              (N02]A = [NO]*omG - [NO]*REM=	  ppm
                                            % Chart
                                      NO2
                                                             'DC
"Calculated concentration from NO or NOX audit calibration equation (y = mx + b).

-------
                          Section 2.0.12
                                                     20
                                                                      Juno 1984
  Tabla 12.4  (continued)

  PART III,  DATA TABULATION

  NO channel
Analyzer— NO
Concentration,
Point Audit concentration, ppm ppm Response
Difference
Analyzer-audit,
ppm %
Zero
10%
20%
40%
60% ' ' ' '
90%
 Slope (m)	;  Intercept (b)



 NOX channel
 .;  Correlation (r)
            Audit concentration, ppm
         Analyzer—NOX
                                                                                    Difference
Point       NO    NOg   NOX total
                                        Concentration,
                                            ppm
                     Response
Analyzer-audit,
     ppm
 Zero
 10%
 20%
40%
60%
90%
Analyzer response (ppm) = m (audit) + b

Slope (m)	;  Intercept (b) =	
.;  Correlation (r)

-------
                             June 1984
                                                      21
                                                                        Section 2.0.12
 Table 12.4  (continued)

 NO2 channel
                                                  Analyzer—NO2
                                                            Difference
 Point       Audit concentration, ppm
                                         Concentration,
                                   Response
Analyzer-audit,
     ppm
Zero
Analyzer response (ppm) =* m (audit) + b

Slope (m) =	;  Intercept (b) =	
              .;  Correlation (r) =.
Converter efficiency
Point number
[NO2]A, ppm
                                                     [NO2]CoNV, ppm
                                                       Percent converter efficiency

-------
                             Section 2.0.12
                                                         22
                                                                           June 1984
    lions in combination with approxi-
    mately 0.10 ppm of NO to challenge
    the calibration of the NO2 channel of
    the analyzer. The NO2 audit concen-
    trations are calculated from the re-
    sponses of the NO channel of the an-
    alyzer using the  NO audit calibration
    equation obtained during the NO-
    NOX audit.
     NO* Audit Procedure—
     1.   Verify that the O3 generator
   air flow rate (F0) is adjusted to the
   value determined earlier (Dynamic
   parameter specifications).
     2.  Generate the SLAMS audit con- •
   centrations (which are compatible
   with the analyzer range) consistent
   with the Appendix A1 requirements.
Audit point
1
2
3
4
Concentration range
(ppm)
0.03-0.08
0.15-0.20
0.35-0.45
0.80-0.90
    3. Generate an NO concentration
  which is approximately 0.08 to 0.12
  ppm higher than the NO2 audit con-
  centration level required. Allow the
  analyzer to sample this concentration
  until a stable  response is obtained;
  that is, until the response does not
  vary more than ±2 % of the
  measurement range over a 5-minute
  period. Record the NO and NOX re- "
  sponses on the data form. Calculate
  and record {NOJ0RiG and [NOX]ORIG
  using the NO  and NOX audit calibra-
  tion equations derived during the
  NO-NOX audit.
   4. Adjust the O3 generator to gen-
  erate sufficient 03 to produce a de-
  crease in the NO concentration
  equivalent to the NO2 audit concen-
  tration level required. After the ana-
  lyzer response stabilizes, record the
  NO and NOX responses on the data
 form. Calculate and record [NOJREM
 and (NO?]REM using the NO and NOX
 audit calibration equations derived
 during the NO-NOX audit.
 {Note: (NO]REM should be approxi-
 mately 0.08 to  0.12 ppm for each
 audit point.)
  5. Calculate  and record the  N02
 audit concentration [NO2JA.
     (N02)A = (NOlORl<3 - [NO]REM
                      Equation 12-13

  6. Obtain the NO2 station response
and concentration from the station
operator and record on the data
form.
  7. Generate the highest  audit con-
centration level first, and consecu-
    tively generate audit points of de-
    creasing N02 concentration. Allow
    the analyzer to sample the audit at-
    mospheres until stable responses are
    obtained. Obtain the necessary data
   • and record in the appropriate spaces
    in Table 12.4.
     8. If desired, additional points  at
   .upscale concentrations different from
   those specified in step 2, may be
   generated. These additional audit
   points plus the zero air point (ob-
   tained at the start of the audit) will
   enhance the statistical significance of
   the audit data regression analysis.
     9. After supplying all audit sample
   concentrations and recording all
 . data, reconnect the analyzer sample
   line to the station manifold. Make a
   notation of the audit stop time. Have
   the station operator make a note on
   the data recorder to indicate the stop
   time, and check all equipment to en-
   sure that it is in order to  resume nor-
   mal monitoring activities.
    Converter efficiency—[Np2]CONV is
   calculated for each audit point using
   Equation 12-14 and is used  to deter-
  mine the NOX analyzer converter effi-
  ciency  using Equation 12-15.
  [NOJoRiG and [NOX]REM are calcu-
  lated from the NOX audit calibration
  equation.
  (N02]CONV = [N02JA - ([NOX]ORIG

             -(NOX]REM)
                      Equation 12-14

  % converter _ [NO2]CONV
   efficiency  ~  [NO2]A  x 10°
                      Equation 12-15

  12.7.6  Calculations—Record the au-
  dit data in the appropriate spaces of
  Table 12.4.
   Percent difference—The % differ-
  ence is calculated as follows:


    % difference =
                     *
                      Equation 12-16
 where
   CM = station-measured concentra-
        tion, ppm,  and
   CA = calculated audit concentration
        ppm.

   Regression analysis—Calculate by
 least squares the slope, intercept,
 and correlation coefficient of the sta-
 tion analyzer response data (y) ver-
 sus the audit concentration data.
 These data can be  used to interpret
 analyzer performance.

 12.7.7  Deference—References 4
through 6, 8, 10, and  12 provide ad-
    ditional information on the N02 audit
    procedure.

    12.8  Carbon Monoxide Audit
    Procedure Using Dynamic Di-
    lution of a Gas Cylinder

    12.8.1   Principle—A dynamic cali-
    bration system used to generate CO
    concentrations for auditing continu-
    ous ambient analyzers, consists of di-
    luting a CO gas cylinder with clean
    dry air.

    12.8.2  Applicability—Dynamic dilu-
   tion can be used to audit all types of
   CO analyzers; CO concentrations in
   the range of 0 to 100 ppm can be
   generated.

   12.8.3  Accuracy—The accuracy of
   the audit procedure should be within
   ±2.5%  if the CO gas cylinder concen-
   tration  is referenced and if gas flow
   rates are determined using recom-
   mended procedures.
   12.8.4  Apparatus—An audit system
   which uses a dynamic dilution device
   to generate audit concentrations is il-
   lustrated in Figure 12.8. The seven
   components of the system are dis-
   cussed below.
    1.  Gas cylinder regulator. A brass
   regulator is acceptable. A low dead-
   space, two-stage regulator should  be
   used to achieve rapid equilibration.
    2.  Flow controllers. Devices capa-
  ble of maintaining constant flow
  rates to within ±2% are required.
  Suitable flow controllers include
  brass micrometering valves in
  tandem with a precision regulator,
  mass flow controllers, capillary re-
  strictors, and porous plug restrictors.
    3. Flowmeters.  Flowmeters capa-
  ble of measuring pollutant and  dilu-
  ent gas flow rates to within ±2% are
  required. NBS-traceable soap bubble
  flowmeters, calibrated mass flow
 controllers  or mass flowmeters, and
 calibrated orifice, capillary, and
 porous plug restrictors.are suitable
   4. Mixing chamber, A glass.or
 Teflon chamber is used to mix the
 CO with dilution air. The inlet and
 outlet should be of sufficient diame-
 ter so  that the chamber is at atmo-
 spheric pressure under normal oper-
 ation, and sufficient turbulence must
 be created in the chamber to facili-
 tate thorough mixing. Chamber vol-
 umes in the range of  100 to 250 cm3
 are sufficient. Glass Kjeldahl connect-
 ing flasks are suitable mixing cham-
 bers.
  5. Output manifold and sample
line. An output manifold used to
supply the analyzer with an audit at-

-------
                                June 1984
                                                         23
                                                                           Section 2.Oil2
                      Extra Outlets Capped
                        When Not in Use
          To Inlet of Analyzer
            Being Audited
   Figure 12.8   Schematic diagram of a dilution audit system.
  mosphere at ambient pressure
  should be of sufficient diameter to
  ensure a minimum pressure drop at
  the analyzer connection, and the
  manifold must be vented so that am-
  bient air will not mix with the audit
  atmosphere during system opera-
  tions. Recommended manifold mate-
  rials are glass or Teflon. The sample
  line must be nonreactive and flex-
  ible; therefore. Teflon tubing is pre-
  ferred.
    6.  Dilution air source. The diluent
  source must be free of CO and water
  vapor. Clean dry air from a com-
  pressed  gas cylinder is suitable
  choices for dilution air. A catalytic
  oxidizer  connected in line is one
  method  of scrubbing  CO from the di-
  lution air.
    7.  CO gas cylinder. A compressed
  gas cylinder containing  100 to 200
  ppm  CO in an air or N2 matrix is
  used as the CO dilution  source. If the
  CO standard is contained in  a N2 ma-
  trix the zero air dilution  ratio cannot
  be less than 100:1. This  cylinder
  must be  traceable to an  NBS-SRM
  (number 1677, 1678 1679, 1680  or
  1681).

  12.8.5 Procedure—Equipment setup
 —Assemble the  audit equipment as
 required, and verify that all the
 equipment is operational. If a clean
 dry air system equipped with a cata-
 lytic oxidizer is used, allow the oxi-
 dizer to warm up for 30 min. Connect
.the gas regulator to the CO cylinder,
land evacuate the regulator as fol-
 lows:
   1. With the cylinder valve closed,
 connect a vacuum pump to the evac-
  uation outlet on the regulator, and
  start the pump.
   2. Open and close the evacuation
  port.
   3. Open  and close  the  cylinder
  valve.
   4. Open and close the evacuation
  port.
   5. Repeat steps 2  through 4 five
  more times  to be sure all O2 impuri-
  ties are removed from the regulator.
  If the regulator does not have an
  evacuation port but has a supported
  diaphragm, the procedure can  be
  conducted at the gas exit port.
   For regulators that do not have an
 evacuation port but have an unsup-
" ported  diaphragm,  use the following
 procedure:
   1. Connect the  regulator to the
 cylinder, and close  the gas exit port.
   2. Open and close the cylinder valve
 to pressurize the regulator.
   3. Open the gas exit port,, and allow
 the gas to purge the regulator.
  4. Repeat  steps 2 and 3 five more
times; then close the gas exit port,
and open the cylinder valve. (The.
regulator should remain  under pres-
sure.) Connect the gas cylinder to the
audit device.
  Repeat the procedure for each
cylinder.
  Audit sequence—After all the
equipment has been assembled and
set up, have the station operator
mark the strip chart recorder to indi-
cate that an audit is beginning. Infor-
mation such as the auditor's name,
start time, date, and auditing organi-
zation should be entered. If it is not
possible to enter written comments.
  the start and stop times should be
  recorded to preclude the use of audit
  data as monitoring data. After
  recording the data, disconnect the
  analyzer sample line from the station
  manifold, and connect it to the audit
  manifold, as shown in Figure 12.9.
  Cap the sample port on  the station
  manifold. The audit atmosphere
  must be introduced through any as-
  sociated filters or sample pretreat-
  ment apparatus to duplicate the path
  taken by an ambient sample. Record
  the analyzer type and  other identifi-
  cation data on the'data form
  (Table 12.5).
    Conduct the audit as follows:
    1. Introduce into the audit manifold
  a  clean dry. air at a flow rate in ex-
  cess of 10% to 50% of the analyzer
  sample demand. Allow the analyzer
  to sample the clean dry air until a
  stable response is obtained;  that is,
  until the response does not vary
  more than ±2% of the measurement
  range over a 5-min period. Obtain
  the station response and concentra-
  tion from the station operator, and
  record the data in the appropriate
  spaces on the data form.
    2. Generate the SLAMS audit con-
  centrations (which are  compatible
  with the analyzer range) as audit at-
  mospheres consistent with the Ap-
  pendix A1 requirements.
  Audit point
       1
       2
       3
       4
 Concentration range,
	(ppm)
        3-8
       15-20
       35-45
       80-90
 Generate the audit concentrations by
 adjusting the pollutant flow rate (FP)
 and the total flow rate (FT) to provide
 the necessary dilution factor.
   Calculate the audit concentration
 as follows.


  [CO] = ^ x [CO]STD  Equation .12-17
 where
     [CO] = audit concentration  of
            CO, ppm
       FP = pollutant flow rate, cm3/
            min
       FT = total flow rate, cm3/min
            [equal to the sum of the
            pollutant flow rate (FP)
            and the dilution flow
            rate (FD)J, and
 [COlsTD = concentration of the
            standard cylinder, ppm.
  3. Generate the highest audit con-
centration level  first, and consecu-

-------
                         Section 2.0.12
                                                    24
                                                                      June 1984
Table 12.5.    Carbon Monoxide Audit Data Report
                                                                                         Date
Address
TA	
                                                                                    Start time
                   °C;  PA
mm Hg;
Analyzer —
Calibration standard
Last calibration date
                                                                Frequency
                                      Serial number
                                      _ Span source
                                      	  Range
Calibration comments
Data acquisition	
Zero setting	
Span setting	
                                                             Data acquisition system
                                                            	^_  Recorder
                                             Bubble flowmeter serial number
Audit standard
    n, dry air _
                                      ;  P
                          ig;  [ ] = •
Flow correction:
Dilution air flow:
    Volume	
                    /PA ~ PH2Q\   f 298 K  \ =
                    \760mm)   \TA + 273J
                                                                       Flowmeter
                                                        mm
                                                                             Volume
Clean, dry air response
    Other response	
                                           % Chart;
                                VDC;
Audit Point I
    Pollutant flow measurement
    Volume	
        T,  	
                                                                       Flowmeter
    Analyzer response _
        Other response
                                           % Chart;
                                                      mm
                            Audit concentration
                            	 VDC;  	
ppm
                                                                                                     No
                                                                                                      cm-3
                                                                                                      min
                                                                                                     ppm
 mn
ppm
ppm

-------
                           June 1984
                                                  25
                                                                   Section 2.O.I2
 Table 12.5  (continued)
 Audit Point II
    Pollutant flow measurement
    Volume
    Analyzer response
        Other response
 Audit Point III
   . Pollutant flow measurement
    Volume   '	.	
        7, 	_
        7i	_
    Analyzer response
       Other response _
Audit Point IV
    Pollutant flow measurement
    Volume
       T2
       T3
   Analyzer response _
       Other response
Audit Point V
    Pollutant flow measurement
    Volume 	
       r,	
       T2 	
       T3 	
   Analyzer response _
       Other response
                                   rn7
                                         %
                                   Cm
                                                    mm
                                         % Chart;
                                   Cm
                                                   mm
                                          Chart:
                                                                    Flowmeter
                                                                      'A  T
                                                                    Audit concentration
                                                                   Flowmeter
                                                                   ff,\jVolume\
                                                                   (Cf:)(—^r~)--
 Audit concentration
	  VDC;  	
                                                                   lr\(yolume\
                                                                   \Cf)\   T  1 ~	
                                                                   Audit concentration
                                                                   Flowmeter
                                                                   (Cf)(^p?)
 Audit concentration
— VDC;  	
                                                                                                  cnr3
                                                                                                  min
                                                                                                 ppm
                                                                                                 ppm
                                                                                                 cmj
                                                                                                 min
                                                                                                ppm
                               cmj
                               min
                                                                                                ppm
 cmj
 min
ppm
Ppm

-------
                          Section 2.0.12
                 28
                                                                        June 1984
 matrix are used as the audit gases.
 These cylinder must be traceable to
 an NBS-SRM (number 1677, 1678,
 1679,1680, or 1681), and must be
 within the following concentration
 ranges: 3 to 8 ppm, 15 to 20 ppm, 35
 to 45 ppm, and 80 to 90 ppm.

 12.9.5  Procedure—Equipment
 setup—Assemble the audit equip-
 ment as required and verify that all
 the equipment is operational. If a
 clean dry air system equipped with a
 catalytic oxidizer is used for a zero
 air source, allow the oxidizer to
 warm up for 30 min. Connect the gas
 regulator to a CO cylinder, and evac-
 ulate the regulator as follows:
   1.  With the cylinder valve closed,
 connect a vacuum pump to the evac-
 uation outlet on the regulator and
 start  the pump.
   2.  Open and close the evacuation
 port.
   3.  Open and close the cylinder
 valve.
   4.  Open and close the evacuation
 port.
   5.  Repeat steps 2 through 4 five
 more times to be sure all 02 impuri-
 ties are removed, from the regulator.
 If the regulator does' not have an
 evacuation port but has a supported
 diaphragm, the procedure can  be
 conducted  at the gas exit port.
   For regulators that do not have an
 evacuation port but have an unsup-
 ported diaphragm, use the following
 procedure:
   1.  Connect the regulator to  the
  cylinder, and close the gas exit
  port.
  2.  Open  and close the cylinder
  valve to pressurize the regulator.
  3.  Open  the gas exit port, and
  allow the gas to purge the regula-
  tor.
  4.  Repeat steps 2 and 3 five more
  times; then close the gas exit port,
  and open the cylinder valve.  (The
  regulator should remain under
  pressure.) Connect the gas cylinder
  to the audit device.
  Repeat the procedure for each
cylinder.
  Audit sequence—After all the
equipment has been assembled and
set up, have the station operator
mark  the strip chart recorder to-indi-
cate that an audit is beginning. Infor-
mation such as the auditor's name,
start time, date, and auditing organi-
zation should be entered. If it is not
possible to enter written comments,
the start and stop times  should be
recorded to preclude the use of audit
 1ata as monitoring data. After
 acording the data, disconnect the
 analyzer sample line from the station
 manifold, and connect it to the audit
 manifold, as shown in F:igure 12.11.
 Cap the sample port on the station
 manifold. The audit atmosphere
 must be introduced through any as-
 sociated filters or sample pretreat-
 ment apparatus to duplicate the path
 taken by an ambient sample. Record
 the analyzer type and other identifi-
 cation data on the data form
 (Table 12.6).
  Conduct the audit as follows:
  1. Introduce  into the audit mani-
 fold a zero air gas at a flow rate in
 excess  of 10% to 50% of the analyzer
 sample demand. Allow the analyzer
 to sample the zero air until a stable
 response is obtained; that is, until
 the response does not vary more
 than ±2% of the measurement range
 over a 5-min period. Obtain the sta-
 tion response and concentration
 from the station operator, and record "
 the data in the appropriate spaces on
 the data form.
  2. Generate the SLAMS audit con-
 centrations (which are  compatible
 with the analyzer range) as audit at-
 mospheres consistent with the Ap-
 pendix A1 requirements.
Audit point
1
2
3
4
Concentration
range,
ppm
3-8
15-20
35-45
80-90
  3. Generate the highest audit con-
centration level first, and consecu-
                  Station Manifold
 tively generate decreasing concentra-
 tions. The audit concentration equals
 the CO gas cylinder concentration.
  4. If desired, additional points at
 upscale concentrations different from
 those specified in step 2 may be gen-
 erated. Generation of these audit
 concentrations plus a post audit
 clean dry air response will enhance
 the statistical significance of the
 audit data regression analysis.
  5. After supplying all audit concen-
 trations and recording all data, re-
 connect the analyzer sample line to
 the station manifold. Make a notation
 of the audit stop time. Have the sta-
 tion operator make a note on the
 data recorder to indicate the 'stop
 time,  and check'all equipment to en-
 sure that it is in order to resume nor-
 mal monitoring activities.

 12.9.6  Calculations—Record the
 data in Table 12.6 in the appropriate
 spaces.
  % difference—The % difference is
 calculated as follows:
    % difference =
                 CM - CA
x 100,
                    CA
                     Equation 12-19
  CM = station-measured  concentra-
       tion, ppm
  CA. = the calculated  audit concen-
       tration, ppm.

  Regression analysis—Calculate by
least squares the slope, intercept,
and correlation coefficient  of the sta-
tion analyzer  response  data (y) ver-
sus the audit  concentration data (x).
These data can be used to  interpret
the analyzer performance.
                    T   T   T  T
                    To Analyzers
     Test
  Atmosphere

1 J— 1
Station
Analyzer-


Data
Acquisition
System
(Teletype
\Printout
H . «
in \
Volts

                      -(Audit Manifold-*- Exhaust
Figure 12.11.    Schematic of configuration utilized in auditing the gas analyzers.

-------
        27
                          Section 2,0.1 Z
 ng the gas analyzers.


 e station analyzer re-
 3 (y) versus the audit con-
 iata (x). These data can
 interpret the analyzer per-
 srences—References 4
 nd 10 and 13 provide ad-
 irmation on the CO audit


 bon Monoxide Audit
 e Using Multiple
 ation Gas Cylinders

 ic/p/e—Separate corn-
 s-cylinders which contain
 concentrations are sup-
 ess to a vented manifold;
 r which is being audited
 ch concentration until a
 3nse results.

 ilicability—The procedure
 I to audit all types  of CO
 oncentrations of CO  in
 f 0 to 100 ppm can be
  12.9.3  Accuracy—The accuracy of
  the audit procedure should be within
  ±2.5% if the CO gas cylinder concen-
  tration is referenced and if gas flow
  rates are determined using EPA-
  recommended procedures.

  12.9.4  Audit Apparatus—A system
  used to generate audit concentra-
  tions is illustrated in Figure 12.10.
  The six components of the system
  are described below.
   1. Gas cylinder regulator. A brass
  cylinder regulator is acceptable.  A
  low deadspace, two-stage regulator
  should be used to achieve rapid
  equilibration.
   2. Flow controllers. Devices capa-
  ble of maintaining constant flow
  rates within ±2% are required. Suit-
  able flow controllers include brass
  micrometering valves in tandem  with
  a precision regulator, mass flow  con-
 trollers, capillary restrictors, and
 porous plug restrictors.     :
   3.  Flowmeters. Flowmeters capa-
 ble of measuring pollutant gas and
 diluent air gas flow rates within ±2%
 are required. NBS:traceable soap
 bubble flowmeters,  calibrated mass
 flow controllers or mass flowmeters,
 and calibrated orifice, capillary, and
 porous plug restrictors are suitable
 for flow determination.
   4. Output manifold and sample
 line. An output manifold is used to
 supply the analyzer with an audit at-
 mosphere at ambient pressure. The
 manifold should be of sufficient di-
 ameter to ensure a minimum pres-
 sure drop at the analyzer connection,
 and the manifold must be vented so
 that ambient air will not mix with the
 audit atmosphere during system op-
 erations. Recommended manifold
 materials are glass or Teflon. The
 sample line  must be nonreactive and
 flexible; therefore. Teflon tubing is
 preferred.
  5.  CO gas cylinders. Compressed
gas cylinders containing CO in an  air
'Under
lulator
      Vent-
        Extra Outlets Capped
          When Not in Use
Output
Manifold
JJ U

-IT'
Fr

                                   To Inlet of Analyzer
                                    Being Audited
   Schematic diagram of a dynamic audit system.

-------
                        Section 2.0.12
                                                   26
                                                                     June 1984
Table 12.5  (continued)
PARTI
Location
Analyzer/model number

Serial number	

Auditor _	
Start time
Zero setting
PART II
                                                                             Date
                                  Pollutant cylinder no.
                                     Pollutant cylinder concentration

                                     Stop time	
                   Span setting
                                                                      Time constant
 Point
number
cm3/min	FT/ cm3/min
    Audit
concentration,
    ppm
                Analyzer
Analyzer      concentration,          %
response	ppm	difference
 Zero
                                Zero
 PART III.  REGRESSION ANALYSIS

 Analyzer response (ppm) = m (audit) + b

 Slope (m)	 ;  Intercept (b)

 Comments:
                                                   ;  Correlation (r) =.

-------
                            Juno 1984
                                                     29
                                                      Section 2.0.12
 Table 12.6.  Carbon Monoxide Audit Data Report

 PARTI
 Location
 Analyzer/model number.

 Serial number	

 Auditor	
 Start time
 Zero setting.

 PART II
                                                                                  Date
                                                  Pollutant cylinder no.
                                                  Pollutant cylinder concentration

                                                  	 Stop time      .
                                           Span setting
                                                  Time constant
 Point
 number
 Audit
cylinder
number
NBS reference
    audit
concentration,
    pom
                                                                Analyzer
                                                     Response
Concentration,
    ppm
    %
difference
 Zero
 Zero
PART III.  REGRESSION ANALYSIS

Analyzer response (ppm) = m (audit) + b

Slope (m) =	; Intercept (b) =	; Correlation (r) =.

Comments:

-------
                          Section 2.0.12
                                                      30
                                                                        June 1984
 12.9.7  References—References 4
 through 6 and 10 and 13 provide ad-
 ditional information on the CO audit
 procedure.

 12.10  Ozone Audit Procedure
 Using Ultraviolet Photometry

 12.10.1 Principle—O3 concentra-
 tions are generated by using a UV
 generator (transfer standard), and
 each atmosphere is verified by using
 UV photometry. The UV photometry
 procedure for 03 audits is based on
 the Lambert-Beer absorption law:

      Transmittance = r- - e"ac1
                     •o
                     Equation 12-20
 where
  a = the absorption cofficient of O3 at
    254 nm =* 308 ±4 atm~1 cm~1  at
    O'C and 760 torr,
  c » the 03 concentration, atm, and
   I = the optical path length, cm.
12.10.2 Applicability—The proce-,
dure can be used to audit all types of
commercially available 03 analyzers
which operate in the range of 0 to
1 ppm.

12.10.3 Accuracy—The accuracy of
the audit procedure should be within
±2.5% if the O3 source is a photome-
ter or transfer standard, and flow
rates are determined using EPA-
recommended procedures.

12.10.4 Apparatus—An UV photo-
metric system which is used for au-
diting O3 analyzers is illustrated in
Figure 12.12. The system  consists of
an O3 source and a standard UV pho-
tometer. Components of the system
are discussed below.
  1.   Ozone generator. An O3 gener-
ator that produces a stable 03 con-
centration is required. An UV lamp
generator is recommended.
  2.   Flow controllers. Devices capa-
ble of maintaining constant flow
             rates to within ±2% are required.
             Suitable flow controllers include
             micrometering valves in tandem with
             a precision regulator, mass flow con-
             trollers, capillary restrictors, and
             porous plug  restrictors.
              3.  Flowmeters. Flowmeters capa-
             ble of measuring clean dry air flow
             rates to within ±2% are required.
             NBS-traceable soap bubble flowme-
             ters, calibrated mass flow controllers
             or mass flowmeters, and calibrated
             orifice, capillary, and porous plug re-
             strictors are suitable.

              4. Mixing chamber. A glass or
             Teflon chamber-is used to mix the O3:
             with dilution  air. The inlet and outlet *
             should be of  sufficient diameter so
             that the chamber is at atmospheric
             pressure under normal operation,
             and sufficient turbulence must be
             created in the chamber to facilitate
             thorough mixing. Chamber volumes
            .in the range of 100 to 500 cm3 are
        \
 Clean
  Dry
  Air
Flow
Controller


-
Flowmeter



Fo .

Fo


             Controller
                                         Generator
                 Mixing
                Chamber
                Flow
              Controller
                                      Output
                                     Manifold
                                                                                                        Vent
 1
                           Extra Outlets Capped
                            When Not in Use
                                                                     To Inlet of Analyzer
                                                                       Being Audited
                           Two-Way
                            Valve
                                                        Vent
              r
                           UV Photometer
                          Detector
                                                     Absorption Cell
                           Signal
                          Processing
                          Electronics
                                          Optics
0                                                  Source

                                                    °
         Flowmeter
  Flow
Controller
Pump
                                                     Exhaust
Figure 12.12   Schematic diagram of an ultraviolet photometric audit system.

-------
                              June 1984
                   31
Section 2.0.12
  sufficient. Glass Kjeldahl connecting
  flasks are suitable mixing chambers.
    5.  Output manifold. An output
  manifold is used to supply the ana-
  lyzer with an audit atmosphere at
  ambient pressure. The manifold
  should be of sufficient diameter to
  ensure a  minimum pressure drop at
  the output ports, and the manifold
  must be vented so that ambient air
  will not mix  with the audit atmo-
  sphere during system operations.
    6. Sample line and connecting
  lines. The sample line and connect-
  ing lines downstream of the 03 gen-
  erator must be made of a  nqnreac-
  tive material such as Teflon.
    7. Dilution air system. Clean dry
  air from a compressed gas cylinder
  (Grade 0.1) is a suitable source of di-
  lution  air;  however, if large volumes
  of air (5 liters/min or greater) are re-
  quired, purified compressed air is
  preferred. The clean dry air must be
  free of contaminants, such as NO,
  N02, O3, or reactive, hydrocarbons.
  The air can be purified to meet these
  specifications by passing it through
  silica gel for drying, by treating it
  with 03 to convert any  NO to NO2,
  and by passing it  through activated
  charcoal (6-14 mesh) and a molecular
  sieve (6-16 mesh, type 4A) to remove
•  N02, 03, or-hydrocarbons.
   Silica gel maintains its drying effi-
  ciency  until it has absorbed 20% of
  its weight; it can be regenerated in-
  definitely at 120°C. Addition of cobalt
 chloride to the surface of the gel pro-
 vides a water absorption indicator. A
 transparent drying column  is recom-
 mended. Activated charcoal and a
 molecular  sieve have a  finite absorp-
 tion capability; because it is difficult
 to  determine when the capability  has
 been exceeded, both should be re-
 placed  either  before each audit  or
 after 8 h of use.
  8. Ultraviolet photometer. The UV
 photometer consists of a low-
 pressure mercury discharge lamp,
 collimator  optics, an absorption cell,
 a detector, and signal-processing
 electronics, as illustrated in Fig-
 ure 12.12. The photometer must be
 capable of  measuring the transmit-
 tance, \l\0, at a wavelength of 254 nm
with sufficient precision for the
standard deviation of the concentra-
tion measurements not to exceed the
greater  of 0.005 ppm or  3% of the
concentration. Because the low-
pressure mercury lamp radiates at
several wavelengths, the photometer
must incorporate suitable means to
be sure  that no O3  is generated in
the  cell by the lamp and that at least
99.5% of the radiation sensed by the
  detector is 254-nm radiation. This
  goal can be achieved by prudent se-
  lection of optical filter and detector
  response characteristics. The length
  of the light path through the absorp-
  tion cell must be known with an ac-
  curacy of at least 99.5%.  In addition,
  the cell and associated plumbing
  must be designed to minimize loss of
  O3 from contact with cell walls and
  gas  handling components.
    9. Barometer. A barometer with an
  accuracy of ±2 torr is required to de-
  termine the absolute cell pressure.
    10.  Temperature indicator. A tem-
  perature indicator accurate to ±1°C is
  required to determine the cell tem-
  perature.
  12.10.5  Procedure
   Equipment setup — Assemble the
  audit equipment according to Fig-
  ure 12.12. Allow the photometer and
 •O3 generator to warm up for approxi-
  mately 1 h or until the normal oper-
  ating cell temperature, 6° to 8°C
  above ambient, is attained.
   Photometer adjustment (Dasibi) —
  Several checks are made after the
  photometer has reached normal- op-
  eration temperature.
   1.  Switch the photometer to sam-
  ple frequency. Using Table 12.7,
  record and calculate the mean of five
 consecutive readouts. The mean
 sample frequency should be between
 45.0  and 49.0.
   2.  Switch the photometer to con-
 trol frequency. Using Table 12.7,
 record and calculate the mean of five
 consecutive readouts. The mean con-
 trol frequency should be between
 23.0 and 28.0.
   3.  Switch the photometer to span.
 Record this span number and calcu-
 late a new span number as follows:
 Span number = 45.684 x
                        V
                          .6\
                     273.16   )'
                     Equation 12-21
where
  Pb = barometric pressure, mm  Hg,
     and
  Tc = cell temperature, °C.

Dial in the new span number on the
photometer, and display the correct
entry.

  4.  Switch the selector to the oper-
ate position, and adjust the flow ro-
tometer to 2 (Vmin. Using the offset
adjust control on the front panel of
the photometer, set the instrument to
  read between 0.005 and 0.010 while
  sampling clean dry air.
    5.  Determine the true zero display
  reading by recording  10 consecutive
  display updates from  the panel
  meter. Calculate the mean of these
  10 readings.
    Audit sequence—1. Adjust the
  clean dry air flow rate through the 03
  generator to meet the range specifi-
  cations of the station analyzer and
  the 03 output capability of the gener-
  ator. Adjust the dilution clean dry-air
  flow rate so that an excess  air flow
  rate of 10-to  50% of the station ana-
  lyzer and  photometer  sample de-
  mand is generated. Mark the  data ac-
  quisition system to indicate that  an
  audit is beginning, and disconnect
  the sample line from the station
  manifold.  Plug the disconnected
  sample port to the station manifold.
   2.  Connect the audit analyzer and
  photometer to the-output manifold,
  as shown  in Figure  12.12. Allow the
  station analyzer and photometer  to
  sample the clean dry air until  a sta-
  ble response is obtained; that is,
  until the response does not  vary
 more than ±2% of the measurement
 range over a  5-min period. Obtain
 the analyzer response  from the sta-
 tion operator, and. record the data
 and the photometer response in the
 appropriate spaces in Table  12.7.
   3.  Generate the following SLAMS
 audit concentrations (which  are com-  .
 patible with the analyzer range) as
 audit atmospheres consistent  with
 the Appendix A1 requirements.
Audit point
1
2
3
4
Concentration
range, ppm
0.03-0.08
0.15-0.20
0.35-0.45
0.80-0.90
 Record ten consecutive display up-
 dates of the photometer for each
 audit point. Calculate and  record the
 mean of these ten updates. Record
 the station analyzer response.  Both
 the photometer and station analyzer
 readings should be taken only after a
 stable response is exhibited by both
 instruments. Calculate the audit con-
 centrations.

    [O3] = RD - Rz,    Equation 12-22
where
  [O3] = the audit concentration of O3,
      ppm,
   RO = the mean of the 10 photo-
      meter display updates, and
   RZ = the average photometer
      clean dry air offset.

-------
                         Section 2.0.12
                32
                                 June 1984
Table 12.7.  Ozone Audit Data Report
Station -
                                                    Date
Address
                                               Start time
 TA
      mm Hg;
 mm Hg  Auditor
Analyzer
 Last calibration date
                            Frequency
         Serial number
        	  Range
 Calibration comments
 Zero setting	
 Span setting	
                        Data acquisition system
                       	  Recorder
Audit system
                Serial number
Clean, dry air supply
Sample frequency —
Control frequency	
              Cell temperature (Tc)
Span number calculation:  45.S84 x (76°pmm} x (7%+7|73) -	
                         Observed span number	
Dilution air:
    Photometer display
    Average
    Analyzer response
      % Chart;'
                                                                                                     ppm
        Other response
Audit Point I
    Photometer display
    Average
         1 setting
    Analyzer
                                   Audit [03]
        Chart;
                             ppm
                             ppm
        Other response
Audit Point II
    Photometer display
    Average
    Sleeve setting
cm
Audit [O3]
    Analyzer response
      % Chart;
                             ppm
                             ppm
       Other response

-------
                             June 1984
                                                     33
                                                                       Section 2.0.12
 Table 12.7  (continued)
 Audit Point III
     Photometer display
     Average
     Sleeve setting
     Analyzer response _
         Other response
 Audit Pomt IV
     Photometer display
     Average
     Sleeve setting
     Analyzer response _
        Other response
 Audit Point V
     Photometer display
    Average
     Sleeve setting
    Analyzer response _
        Other response
AUDIT RESULTS
                                     cm
                                           % Chart; •
 Audit [O3]
- VDC;  _
ppm
ppm
                                     cm
                                                                        Audit [O3]
                                           % Chart;
                              ppm
                              ppm.
                                     cm
                                                                       Audit [O3]
                                           % Chart;
                              ppm
                              ppm
 Point
number
                                                               Analyzer
                    Audit concentration.
                           ppm
                                                 Response
Concentration
    ppm
                                                                                                difference
  1
  2
  3
  4.
  5
  6
   Regression analysis (y = mx + b)
Analyzer response (ppm) = m (audit) + b
             Slope (m) =	
           Intercept (b) =	
          Correlation (r) =	

-------
                          Section 2.0.12
                                 34
                                                                         June 1984
   4. Generate the highest audit con-
 centration level first by adjusting the
 O3 output of the generator, the
 amount of dilution air, or the amount
 of clean dry air flowing through the
 generator. Then consecutively gener-
 ate the decreasing concentrations.
   5. If desired, additional points at
 upscale concentrations different from
 those specified in step 3 may be gen-
 erated. Generation of these audit
 concentrations plus a post audit
 clean dry air response will enhance
 the statistical significance of the
 audit data regression analysis.  ,
   6. After supplying all audit concen-
 trations and recording all data, re-
 connect the  analyzer  sample line to
 the station manifold.  Make a notation
 of the audit stop time. Have the sta-
 tion operator make a  note on the
 data recorder to indicate the stop
 time, and check all equipment to en-
 sure that it is in order to resume nor-
 mal monitoring activities.

 12.10.6  Calculations—Record the
 data in Table 12.7 in the  appropriate
 spaces.
   % difference—The % difference is
 calculate as follows:
    % difference


 where
                  CM - CA
    x 100,
Equation 12-23
  CM 3 the station-measured concen-
     tration, ppm, and
  CA « the calculated audit concentra-
     tion, ppm.

  Regression analysis — Calculate by
 least squares the slope, intercept,
 and correlation coefficient of the sta-
 tion analyzer response data (y) ver-
 sus the audit concentration data (x).
 These  data can be used to interpret
 the analyzer performance.

 12.10.7  References— References 7,
 10, and 14 provide additional infor-
 mation on the O3 audit procedure.

 12.11   Total Suspended Par-
ticulate  Sampler Audit Proce-
 dure Using a Reference Flow
 Device (ReF)
        Principle — An ReF device is
one type of orifice transfer standard
and is used to audit a TSP hi-vol
sampler. The ReF device (Figure
12.13} uses orifice plates to audit the
sampler flow rate by measuring the
pressure drop caused by the flow of
air through a restricting orifice. A
calibration equation is used to trans-
late this pressure drop into a flow
                 Figure 12.13.    Reference flow (Ref) audit device.
 rate at either standard or actual con-
 ditions.

 12.11.2  Applicability—The proce-
 dure can be used to audit hi-vol sam-
 plers with or without flow controllers
 operating in the flow range of 0.5 to
 2.4 std m3/min. Other types of orifice
 transfer standards may be  used fol-
 lowing the same procedures.

 12.11.3  Accuracy—The accuracy of
 the audit procedure is approximately
 2% when traceability is established
 by calibrating the ReF device to a
 Rootsmeter or other primary volume
 measurement device.-

 12.11.4 ' Apparatus—ReF device—
 An ReF device is an interfacing unit
 that attaches to the filter holder of a
 TSP hi-vol sampler. The device typi-
 cally exhibits a sensitivity of 0.01 m3/
 min per 0.1-in. pressure change. The
 ReF device is equipped with five air-
 restricting orifice plates which  are
 used one at a time to vary the  flow
 rate of the hi-vol sampler. A slack
tube water manometer accompanies
the ReF device and measures the
pressure drop caused  by the flow re-
striction of the plates.  A cylindrical
plexiglass windflow deflector should
be attached to the top of the ReF de-
vice to protect it from  ambient air
flow.
  Differential .manometer—A tube
manometer capable of measuring at
least 16 in. of water is required.
   Barometer—A barometer capable
 of measuring atmospheric pressure
 with an accuracy of ±2 torr is re-
 quired.
   Temperature indicator—An indica-
 tor accurate to' ±1°C is required to.
 determine ambient temperature'.
 .  Glass fiber filter—Glass fiber filters
 with at least 99% efficiency for col-
 lection of 0.3-p.m diameter  particl.es
 are suitable.

 12.11.5  Procedure—Samplers
 equipped with flow controllers—A
 hi-vol sampler equipped with a flow
 controller is typically calibrated in
 terms of standard flow rate. Audit
 calculations are performed  as shown
 in Section 12.11.6.
   Note:   It is imperative to know
 whether the hi-vol  was calibrated in
 terms of actual conditions at the time
 of calibration, seasonal average con-
 ditions, or the flow rates have been
 corrected to standard temperature
 and pressure. The comparison be-
 tween  audit and station flow rates
 MUST be made with the same units
 and corrections.
  Conduct the audit as follows:
  1.  Remove the filter holder clamp
from the  sampler. If a filter  is in
place for an upcoming  sampling
period, have the station operator re-
move the filter and store it until the
audit is completed. Attempt to sched-
ule audits so they do not interfere
with normal sampling runs

-------
                              June 1984
                 35
Section 2.0.12
   2. Place a clean glass fiber filter on
  the filter screen, and place the ReF
  device on top of the filter. Securely
  fasten the ReF device to the holder
  using the four wingnuts at each cor-
  ner of the sampler filter holder.
   3. With no  resistance plate in the
  ReF device, close the  lid and fasten it
  using the two wingnuts. Place the
  wind deflector in position, and then
  connect and zero the water manome-
  ter.   .
   4. Start the sampler motor and
  allow it to stabilize. A warm-up time
 •of .2:5 min should be allowed. Record
  the pressure drop shown on the
  manometer (in. H2O), ambient tem-
  perature (°C), barometric pressure
  (mm Hg), and station  flow rate (ob-
  tained from the station operator) on
  the data form  in Table 12.8.  If the
  barometric pressure cannot be deter-
  mined by an audit barometer (be-
  cause of high  elevations that exceed
  the limits of the  barometer), deter-
  mine the barometric pressure (PA) as
  follows:

   PA = 760 - (elevation in meters

        x 0.076).       Equation 12-24


   5.  At the conclusion  of the audit,
  have the  station  operator replace the .
  filter and reset; the sampler timer as
  it was before the audit.
   Samplers without flow controllers
 —A hi-vol sampler not equipped with
  a constant flow controller is typically
  calibrated in terms of  actual flow
  rates. Audit calculations  are per-
 formed as shown in Subsection
.  12.11.6.
   Note:   It is imperative to  know
 whether the hi-vol was calibrated, in  '
 terms of actual conditions at the time.
 of calibration, seasonal average con-
 ditions, or the  flow rates have been
 corrected to standard temperature
 and pressure. The comparison be-
 tween audit and station flow rates
 MUST be made with trie same units
 and corrections.
   Conduct the audit as. follows.:
   1. Remove  the filter holder clamp
 from the sampler. If a  filter is in
 place for an upcoming sampling pe-
 riod, have the station operator re-
 move the filter and store it until the
 audit is completed. Attempt to sched-
 ule audits so they do not interfere
 with normal sampling  runs.
   2. Place the ReF device on  the filter
 holder, and secure the device to the
 holder by tightening the four
 wingnuts  at each corner of the sam-
 ple filter holder.
   3. Place the 18-hole resistance plate
 in the ReF device, close the lid, and
 fasten the lid using the two
 wingnuts. Place the wind deflector in
 position, and then connect and zero
 the water manometer.
   4. Start the sampler motor and
 allow it to stabilize. A warm-up time
 of 2:5 min should be allowed. Record
 the pressure drop shown on the
 manometer (in.  H20), ambient tem-
 perature (°C), barometric pressure
 (mm Hg), and station flow  rate (ob-
 tained from the  station operator) on
 the data form in Table 12.8. If the
 barometric pressure cannot be deter-
 mined by an audit barometer (be-
 cause of high elevations that exceed
 the limits of the barometer), deter-
 mine the barometric pressure by
 using Equation 12-24.
   5. Repeat steps 3 and 4 using the
 remaining resistance plates.
   6. At the conclusion of the audit,
 have the station operator replace t^e
 filter and reset the sampler timer as
 it was before the audit.

 12.11.6  Calculations—Calculate the
 audit flow rate at standard  conditions
 for those hi-vols with flow  rates cor-
 rected to standard temperature and
 pressure.
    For samplers calibrated in terms of
  actual or seasonal average condi-
  tions — Calculate the audit flow rate
  in terms of actual conditions.
                  /760
                   Pb\/298\  ..
                        --
                     Equation 12-25 .
where
      Qstd = standard flow rate,
           m3/min
  m and b = calibration  coefficients
           determined during cali-
           bration of the ReF device,
           using flow rates corrected
           to standard conditions
       AH = pressure drop shown on
           the manometer, in. H2O .
        Pb = barometric pressure,
           mm Hg, and
        Ta = ambient  temperature in
           degrees Kelvin (273.16
  Perform this calculation for each
flow rate comparison and calculate
the % difference for each audit point
as follows.


    % difference = -^-= — - x 100,
                    r~A
                     Equation 12-26
where
  FS = the station-measured flow
     rate, std m-Vmin, and
  FA = the audit flow rate, std m3/min.
                       Equation 12-27
  where
    Qact = the actual flow rate, m3/min
    Qstd = the standard flow rate,
        m3/min
     Pb = the barometric pressure, mm
        Hg, and
     Ta = the ambient temperature in
        degrees Kelvin (273.16 + °C). -

  Note:  If seasonal temperature and
        barometric pressure were used
        in the calibration of the hi-vol
        sampler, then:

     Pb= seasonal barometric pres-
        sure, mm Hg, and
     Ta = seasonal ambient tempera-
        ture in degrees Kelvin (273.16
        + °C)

  convert from m3/min to ft3/min by
  multiplying by 35.31.

  12.11.7  References — References 8
  and 9 provide additional information
  on the TSP audit procedure.

  12.12   Data Interpretation
    Interpretation of quality assurance
  audit results is not well defined, and
  audit data must be assembled and
  presented so that interpretation is
  possible.  Subsection 12.12.1 dis-
  cusses the data reporting require-
  ments specified in Appendix A1. In
  addition to these requirements, op-
  tional data interpretation methods,
  including case .examples, are in Sub-
  section 12.12.2.

  1 2. 1 2. 1   SLAMS Reporting Require-
  ments — Refrence 1 specifies the min-
  imum data reporting procedures for
  automated and manual methods.
  Compare the station responses ob-
  tained for each audit point.
                f^   _ s*
  % difference = — ^~ — - x 100
                  t-A
                      Equation 12-29
  where
    CM  = the station-measured re-
       sponse, in concentration units,
       and
    CA  = the audit values, in concentra-
       tion units.

  This comparison indicates the % dif-
  ference for each audit concentration
  generated and each analyzer re-
  sponse recorded.

-------
                        Section 2.0.12
                                                   36
                                                                    June 1984
Table 12.8.  Hi-vol Sampler Audit Data Report
Station location
Data	
Time	
 Barometric pressure
	  Temperature —
Sampler model number
Flow controller number .
                                                         Serial number
 Plate           Audit manometer      '   Audit
number  '       reading, in. H2O  	flow
                                                            Sampler
                                                                                          Difference
        Response
Flow
                                      m3/min
No plate
   18
   13
   10
Audit device ID number
Other information:
                                                 Qstd; Slope (m) =.
                                                 Qact; Slope (m) =.
                                       Regression coefficient
                                ;  Intercept (b) = .
                                ;  Intercept (b) = .
Audited by:
      Authorized by:

-------
                             June 1984
                                                      37
                                                                         Section 2.0.12
Table 12.9.  Example Audit Data for an SO2 Analyzer
SLAMS
concentration
range, ppm
0.03 to 0.08
0.1 5 to 0.20
0.35 to 0.45
Audit
concentration,
ppm
0.044
0.165
0.412
Station
analyzer
response, ppm
0.042
0.159
0.394
%
difference
-4.6
-3.6
-4.4
  Table 12.9 contains example audit
data for an S02 analyzer operating
on a 0- to 0.5-ppm range. As indi-
cated by the data set, the station an-
alyzer shows a negative deviation of
approximately 4% when compared to
the audit concentrations.
  A % difference calculation is used
to evaluate manual  method audit
data. For example, a hi-vol sampler
with a flow controller is audited
using an ReF device. A one-point
audit is performed at the normal op-
erating flow rate with a glass fiber fil-
ter on the device. The audit and sta-
tion flow rates are compared on the
basis of % difference using Equation
12-29 and are  designated as CA and
CM, respectively.

12.12.2  Least Squares—The data
analysis described in Appendix A1
calculates the  % accuracy of the
audit data at specific operating levels
within an analyzer's range. Because.
this method compares the operating
differences at  a maximum of four
points, its use in determining overall
analyzer performance is limited.
  With an increase in the number
and range of audit points generated,
linear regression analysis can be
used to aid in  evaluating analyzer
performance data. This method in-
volves supplying a zero concentra-
tion and five upscale concentrations
corresponding to approximately 10%,
20%, 40%, 60%, and 90% of the ana-
lytical  range. The  regression coeffi-
cients  are calculated by designating
the audit concentration (ppm) as the
abscissa (x variable) and the station
analyzer response (ppm) as the ordi-
nate (y. variable). The resultant
straight line (y = mx + b) minimizes
the sum of the squares of the devia-
tions of the data points from the  line.
  Table 12.10 summarizes the calcu-
lations by the method of least
squares, and Table 12.11  lists criteria
which  may be used  to evaluate the
regression data in terms of analyzer
performance. The slope and intercept
describe the data set when  fitted to a
line;  the correlation  coefficient de-
scribes how well the straight line fits
the data points. Presently, there are
no published criteria for judging ana-
lyzer performance. Criteria are nor-
mally specified by the operating
agency. Figure 12.14 shows an exam-
pie audit data set that is analyzed
both by the % difference and least
squares technique. The slope shows
an average difference of -4.2%
which agrees with the % difference
data. The zero intercept of 0.000
agrees with the analyzer response
during the audit; this indicates a
nonbias response. The correlation •
                       coefficient of 0.9999 indicates a linear
                       response to the audit points. It can
                       be deduced that the % difference of
                       the slope index is caused by the cali-
                       bration source (i.e., the standard pol-
                       lutant source, flow measurement ap-
                       paratus, and the dilution air source).
                         Figure 12.15 illustrates data varia-
                       tions which may be encountered
                       when auditing a monitored network.
                       Figure 12.15(a) represents audit re-
                       sults in which the analyzer response
                       agrees perfectly with the generated
                       audit concentrations. Figure 12.15(b)
                       represents data from a group of sta-
                       tions showing constant systematic
                       differences, (i-.e., differences inde-
                      'pendent of concentration levels be-  -
                       tween stations and between stations
                       and the audit  system).
                         A network of stations showing lin-
                       ear systematic differences that may
                       or may not be independent of con-
                       centration is shown in Figure
Point No.
1
2
3
4
5
6
Audit .
Concentration
(ppm/
.000
.044
.103
.165
.294
.412 •
Station
Concentration
(ppm)
.000
.042
.098
.159
.283
.394
% Difference

-4.6
-4.9
-3.6
-3.7
-4.4
                                        r  = 0.9999
                                        m = 0.958
                                        b  = 0.000
Figure 12.14.
            .2         .3

        Audit Concentration (ppm)

Example of audit data regression analysis.

-------
                            Section 2.0.12
                                                          38
                                                                             June 1984
  Table 12.10.  Least Squares Calcula-
  tions' (y ~ mx + b)

        —                .     Vv
        x « average x value = =£

        7 - average y value ~ ~j
        slope « m
                             N
    intercept - b - y - nix

correlation coefficient = r = -
      SZY = variance of the y values
              'Sv*   _
           -[N

     S2X « variance of the x values
                                 N
  aFor convenience Sx ;s used for Zx,-

  (and similarly for other sums):


  12.15{c). This example is moie repre-
 sentative of audit data resulting from
 a network of stations. Figure 12.15(d)
 and 12.15(e)  illustrates two special
 cases of the general case shown in
 Figure 12.15(c).
   Analysis of the data for a grouping
 of stations, such as for a given State,
 not only yields precision and accu-
 racy estimates but may also provide
 clues as to the proper corrective ac-
 tion to take if larger than acceptable
 differences are observed. For exam-
 ple. Figure 12.15(d) shows  constant
 relative differences within stations
 that vary among stations. Such data
 patterns can result, for example,
 from errors in the calibration stan-
 dards if high concentration cylinders
 and dilution are used for calibration.
 Constant systematic (absolute) differ-
 ences {within stations), such as Fig-
 ure 12,15(b), may indicate contami-
 nated zero and dilution air, in which
 case all results would tend  to be on
 one side of the 45' line.
  Figure 12.15{e) illustrates a case in
 which stations were calibrated using
 a high concentration span level, but
 not multipoint concentrations or zero
 point.
  The usfe of regression analysis is
not as straightforward when the in-
tercept is significantly different from
                                     a
                                                 Audit Concentration

                                          (a) Audit data from an ideal station.
                                                                                       Audit Concentration
                                                                                    (b) Systematic differences between
                                                                                       station values and audit
                                                                                       values.
                                                                Audit Concentration
                                                          (c) Linear and systematic differences
                                                            between station values and audit
                                                            values.
                                               §
                                               CJ
                                               §
                                               o
                                                  Audit Concentration
                                                                           Audit Concentration
                                                   (d). (e) Differences resulting from inaccurate
                                                         calibration standard.
                                   Figure 12.15.   Multiple audit data variations.
                                  zero and/or the correlation is low
                                  (<0.995). In these instances, the audi-
                                  tor must rely on his experience to
                                  draw conclusions about the cause of
                                  a high or low intercept, a low corre-
                                  lation, and the  subsequent meaning
                                  of the results. The five most com-
                                  monly encountered audit cases are
                                  discussed in the following subsec-
                                  tions.
                                    Case 7 — The data  set and data plot
                                  in  Figure 12.16  illustrates  a case in
                                  which the % difference and the linear
                                  regression analysis of audit data
                                  must be used jointly to characterize
                                  analyzer performance. .Inspection of
the % difference for each audit point
shows large negative differences at
the low concentrations and small dif-
ferences at the upper concentrations.
The slope of the regression line indi-
cates an overall slope  of +2.2% and
a significant intercept of -0.014. The
following statements apply to the re-
gression data.
  1. Analyzer zero drift may have oc-
  curred.
  2. The dilution air source used to
  calibrate  the analyzer has a bias
  (not of sufficient purity).
  3. The calibration procedure  used
  by the operator is not correct.

-------
                             June 1984
                                                       39
Section 2.0.12
Table 12.11.    Linear Regression Criteria
Slope
    Excellent

    Satisfactory     ±6% -

    Unsatisfactory
Intercept
    Satisfactory

    Unsatisfactory
                            s  ± 5%  between analyzer response and audit
                                        concentration
                               ± 75%  between analyzer response and audit
                                        concentration
                            ~>  ± 15%  between analyzer response and audit
                                        concentration
                            s  ±3%   of the analyzer range

                            >  ±3%   of the analyzer range
Correlation coefficient
    Satisfactory   '  0.9950 to (1.0000)  linear analyzer response to audit
                                        concentrations
    Unsatisfactory   <0.9950           nonlinear analyzer response to audit
                                        concentrations


Point No.
1
2
3
4
5
6 -- -
Audit
Concentration
fppmj
.000
.053
.119
.222
.269
.396
Station
Concentration
Ippm)
-.0/3
.043
.703
.203
. .263
.392


% Difference
—
-18.9
-13.5
- 6.3
- 2.2
• 1.0
     .5
     .4
 I
     .3
 §
 
-------
                           Section 2.0.12
                                                       40
                                                                         June 1984


Point No.
1
2
3
4
S
6
Audit
Concentration
(ppm)
.000
.053
.119
.222
.269
.396
' Station
Concentration
(ppm)
.000
.043
.103
.208
.263
.392


% Difference
_
-18.9
-13.S
- 6.3
• 2.2
• 1.0
 o
 1
 8
 I
     .4
      .3
,2
     .1
         Without Zero Incarcept
           f  * 0.3396
           m = 1.026
           b  = 0.0/5
With Zero
Incercapt
  r  = 0.9991
  m = 1.001
  b  = 0.009
                             _L
                                         _L
Flgurt 12.17.
             •I          .2           .3

                     Audit Concentration (ppm)'

          Audit data interpretation—Case 2.
                                                    .4
                                                               .6
 and possibly a significant zero inter-
 cept. A graphic plot will verify sus-
 pected analyzer nonlinearity.
  Case 5—The data illustrated in Fig-
 ure 12.20 show the results of an
 audit performed on a NOX analyzer.
 The regression coefficients show an
 overall difference between the audit
 concentrations and analyzer re-
 sponses of -20.0% and an intercept
 of 0.011 ppm. The analyzer response
 for the zero concentration and first
 four audit concentrations shows a
 constant bias which would be ex-
 pected for the entire range. Percent
 differences for the three remaining
audit levels become increasingly
large. A graphic plot of the audit data
indicates the analyzer converter effi-
                                ciency is decreasing with increasing
                                audit concentration.

                                12.13   References
                                 1. 40 CFR 58, Appendix A—Quality
                                   Assurance Requirements for
                                   State and Local Air Monitoring
                                   Stations (SLAMS), Ambient Air
                                   Quality Surveillance.
                                 2. Ref. 1. July 1, 1984.
                                 3. 40 CFR  58, Appendix B—Quality
                                   Assurance Requirements for Pre-
                                   vention of Significant Deteriora-
                                   tion (PSD) Air Monitoring.
                                 4. Traceability Protocol for Establish-
                                   ing True Concentrations of Gases
                                   Used for Calibration and Audits
                                   of Air Pollution Analyzers, (Proto-
                                   col No. 2). June 15, 1978. Avail-
      able from the U.S. Environmental
      Protection Agency, Environmen-
      tal Monitoring Systems Labora-
      tory, Quality Assurance Branch
      (MD-77), Research Triangle Park,
      NC.
    5. Protocol for  Establishing Trace-
      ability of Calibration Gases Used
      With Continuous Source Emis-
      sion Monitors. August 25, 1977.
      Available from the U.S. Environ-
      mental Protection Agency, Envi-
      ronmental Monitoring Systems
      Laboratory, Quality Assurance
      Branch,  (MD-77), Research Trian-
      gle Park, NC.
   6. Catalog  of  NBS  Standard Refer-
      ence Materials. NBS Special Pub-
      lication 260, U.S. Department of
      Commerce, National Bureau of
      Standards, Washington, DC.
      1984-85  Edition.

   7. Transfer Standards for Calibration
     of Air Monitoring Analyzers for
     Ozone. Technical Assistance Doc-
     ument. EPA-600/4-79-056, Envi-
     ronmental Monitoring Systems
     Laboratory, U.S. Environmental
     Protection Agency, Research Tri-
     angle Park, NC. September 1979.
   8. Quality Assurance Handbook for
     Air Pollution Measurement Sys-
     tems, Volume II—Ambient Air
     Specific Methods. EPA-600/4-77-
     0273, Environmental Monitoring
     Systems  Laboratory, U.S. Envi-
     ronmental Protection'Agency,
     Research Triangle Park, NC.
  9.' Investigation of Flow Rate Calibra-
     tion Procedures Associated with
     the High Volume Method for De-
     termination of Suspended Partic-
     ulates. EPA-600/4-78-047, Envi-
     ronmental Monitoring Systems
     Laboratory, U.S. Environmental
     Protection Agency, Research Tri-
     angle Park, NC. August 1978.
 10.  List of Designated  Reference and
     Equivalent Methods. Available
    from the U.S. Environmental Pro-
    tection Agency, Office of Re-   ,
    search and Development, Envi-
    ronmental Monitoring Systems
    Laboratory, Research Triangle
    Park, NC.
 11. Use of the Flame Photometric
    Detector Method for Measure-
    ment of Sulfur Dioxide in Ambi-
    ent Air. Technical Assistance
    Document. EPA-600/4-78-024,
    U.S. Environmental Protection
    Agency, Environmental Monitor-
    ing Systems Laboratory, Re-
    search Triangle Park, NC.
    May 1978.
12.  Technical Assistance Document
    for the Chemiluminescence Mea-

-------
                           June 1984
41
Section 2.0.12
    surement of Nitrogen Dioxide.
    EPA-600/4-75-003, Office of Re-
    search and Development, Envi-
    ronmental Monitoring Systems
    Laboratory, U.S. Environmental
    Protection Agency, Research Tri-
    angle Park, NC. December 1975.
13.  Guidelines for Development of a
    Quality Assurance Program—
    Reference Method for the Contin-
    uous Measurement of Carbon
    Monoxide in the Atmosphere.
    EPA-R4-73-028a, Office of Re-
    search and Monitoring, U.S. En-
    vironmental Protection Agency,
    Washington, DC. June 1973.
14.  Technical Assistance Document
    for the Calibration of Ambient
    Ozone Monitors. EPA-600/4-79-
    057, Environmental Monitoring
    Systems Laboratory, U.S. Envi-
    ronmental Protection Agency,
    Research Triangle Park, NC. Sep-
    tember 1979.


Point No.
1
2
3
4
S
6
Audit
Concentration
(ppm)
.000
.056
.116
.221
.276
.405
Station
Concentration
(ppm)
•£13
.064
.132
.235
.282
.409


% Difference
	
14.3
13.8
6.3
2.2
1.0
                                                                        r  = 0.9997
                                                                        m = 0.980
                                                                        b  = 0.014
                                                               .2         .3
                                                            Audit Concentration (ppm)
                                   Figurt 12.13.    Audit data interpretation—Case 3.

-------
                              Section 2.0.12
                                                             42
                                                                 June 1984


Point No.
f
2
3
4
5
6
Audit
Concentration
(ppm)
.000
.072
.114
.183
.332
.474
Station
Concentration
(ppm)
.000
.054
.080
.134
.296
.503


% Difference
_
-25.0
-29.8
-26.8
-10.8
+ 6.2
                                                              r  = 0.9899
                                                              m = 1.056
                                                              b  = 0.029
         0&-
                        I
Figura 12,19.
      ./           .2           .3

               Audit Concentration (ppm)

Audit data interpretation—Case 4.
                                                             .4
                                                                          .5

-------
                                  June 1984
                                                               43
                                                                                    Section 2.O.I2


Point No.
1
2
3
4
5
6
7
8
Audit
Concentration
(pom)
.000
.056
.106
.206
.313
.417
.651
.885
Station
Concentration
Ippm)
.000
.049
.094
.180
.273
.355
.540
.703


% Difference

-12.5
-11.3
-12.6
-12.8
-14.9
-17.1
-19.7
       1.0,
   I
   c
   o
   I
   I
        .8 -
.6 -
        .4  -
                                                   r  = 0.9991
                                                   m = 0.800
                                                   b  = 0.011
          0           .2          .4           .6
                             Audit Concentration (ppmj

Figure 12.20.    Audit data interpretation—Case 5.
                                                  .8
                                                               1.0

-------

-------

-------
           	ilia*!.,"'' vu,1.'/ij
                                                iiihr ,	i: iVjIUit 1
                    > ! n».-nfj:.n  .•'':
                                       ".'.i.flE!W	S'BH'liJ'.,	iH'i'lHI
  il	>>: Hrl.'liii.ViJ	IK,-	iilili' 1"'
 ll'i'i'll	B!,i,: •iS-.i'Hri'MSIiBI';-1!;
mwrn,';	ist^«»^
 1 <'n	.<: IBiip,«!'"! "I mlliiliii f: !i, liJiD1"! .: ii! -'i,
                  .,'/ r'l'lllil'1! ' .Hii'

                                              >iiilliiii'"i>vi,:i>>l:i iii'i
                                                ^-i'i
                                                 \J1B5H11

-------
                 Jan.1983
                                                  Section 2.2.0
                        United States
                        Environmental Protection
                        Agency
                          Environmental Monitoring Systems
                          Laboratory
                          Research Triangle Park NC 27711
                        Research and Development
                          EPA-600/4-77-027a
&EPA
Test  Method
                        Section 2.2
                        Reference  Method for the
                        Determination of  Suspended
                        Particulates
                        in  the Atmosphere
                        (High-Volume Method)
                                             Outline
                               Section
                       Summary
                       Method QA Highlights
                       Method Description
                         1. Procurement of Equipment
                           and Supplies
                         2. Calibration of Equipment
                         3. Filter Selection and
                           Preparation
                         4. Sampling Procedure
                         5. Analysis of Samples
                         6. Calculations and Data
                           Reporting
                         7. Maintenance
                         8. Auditing Procedure
                         9. Assessment of Monitoring
                           Data for Precision and
                           Accuracy
                        10. Recommended Standards for
                           Establishing Traceability
                        11. Reference Method
                        12. References
                        13. Data Forms


                       Summary
                        Ambient air drawn into a covered
                       housing and through a filter by a
                       high-flow-rate blower at 1.1 to 1.7
                       mVmin (39 to 60 ftVmin) allows total
                                         Number of
                             Documentation  Pages
                                 2.2        '   !
                                 2.2           T
                                 2.2.1

                                 2.2.2
                                 2.2.3

                                 2.2.4
                                 2.2.5
                                 2.2.6

                                 2.2.7
                                 2.2.8
                                 2.2.9
                                 2.2.10

                                 2.2.11
                                 2.2.12
                                 2.2.13
13
 4

 8
 1
 2

 2
 4
 1
 1

10
 1
10
                         suspended particulates (TSP) in sizes
                         up to 25 to 50 ^m (aerodynamic
                         diameter) to collect on the filter
                         surface. When operated within this
                         range, the high-volume sampler is

-------
                         Section 2.2.0
                                                                           Jan. 1983
 capable of collecting TSP samples for
 24«hour TSP concentrations ranging
 from 2 to 750 //g/std m3. The mass
 concentration (/ug/m3)* in ambient air
 is computed by measuring both the
 mass of TSP collected and the
 (standard) volume of air sampled.
   This method provides a
 measurement of the mass
 concentration of total suspended
 paniculate matter (TSP) in ambient air
 for determination of compliance with
 the primary and secondary National
 Ambient Air Quality Standards for
 Paniculate Matter as specified in
 §50.6 and §50.7 of the Code of
 Federal Regulations. Title 40. The
 measurement process is
 nondestructive, and the size of the
 sample collected is usually adequate
 for subsequent chemical analysis.
 Based on collaborative testing, the
 relative standard deviation (coefficient
 of variation) for single analyst
 precision (repeatability) of the method
 is 3,0 percent. The corresponding
 value for interlaboratory precision
 (reproducibility) is 3.7  percent.
   The absolute accuracy of the
 method is undefined because of the
 complex nature of atmospheric
 paniculate matter and the difficulty in
 determining the "true" paniculate   .
 matter concentration.
   This reference method appears in
 Title  40 of the Code of Federal
 Regulations. Part 50. Appendix B (as
 amended on December 6, 1982, (47
 FR 54912),  A complete copy of the
 Reference Method is reproduced in
 Section 2,2 11.
 Method QA Highlights
  In this quality assurance document
 for the TSP Reference Method (high-
 volume sampler method), the
 procedures are designed to serve as
 guidelines for the development of
 agency quality assurance programs.
 Because recordkeeping is a critical
 part of quality assurance activities.
 several data forms are included to aid
 in the documentation of necessary
 data The blank data forms (Section
 2,2,13) may be used as they are, or
 they may serve as guidelines for
 preparing forms more appropriate to
 the individual agency; partially filled-
 m forms are interspersed throughout
 the text to illustrate their uses.
Activity matrices at the end of
 pertinent sections provide a review of
'AHhouijn TSP is measured in microqrams per
ftanrltVd cubic meter me "standard" is
'lommonly omitted when reporting TSP
          s. ay convention, ;jq/mj for TSP is
        to mean pg  std m'
the material covered in the text
sections. The material covered in this
section for the TSP'method is briefly
summarized here.

1.  Procurement of Equipment
Section 2.2.1 describes the selection
of equipment and the recommended
procurement and calibration checks
for the equipment. It also identifies
the sections of this part of the
Handbook that pertain to specific
equipment and supplies. Figure 1.1
provides an example of a permanent
procurement record.

2.  Calibration of Equipment Section
2.2.2 provides detailed calibration
procedures for the analytical balance,
the relative humidity indicator, the
elapsed-time meter, the- flow-rate
transfer standard, and the high-
volume sampler. This section can be
removed (along with  the
corresponding sections for the other
methods of this volume of the
Handbook) to serve as a calibration
handbook. Table 2.2  at the end of the
Section summarizes  the acceptance
limits for equipment  calibration.

3.  Filter Selection and Preparation
Section 2.2.3 presents important
considerations for the selection,
identification, equilibration, weighing
check, and handling of filters. The
spectro-quality grade filter is
recommended for use when additional
chemical analyses are anticipated.

4.  Sampling Procedure Section 2.2.4
details procedures for filter
installation, performance of
operational checks, sample handling,
and data documentation. Several
photographs are provided to clarify
the installation procedure. Complete
documentation of background
information during the sampling is
one of several quality assurance
activities that are important to future
data validation; particularly important
are any unusual  conditions existing
during collection of the sample. Any
such conditions should be noted.

5.  Analysis of Samples Section 2.2.5
briefly describes  verification of data
from the field, sample inspection,
filter equilibration, and the gravimetric
analysis procedure. The analytical
balance must be checked. The filter
must be equilibrated in a controlled
environment.

6.  Calculation and Data Reporting
Section 2.2.6 describes those
activities pertaining to data
calculations and  reporting. The final
data review, the  data edit or
validation, and the use of standardized
reporting procedures are all important
parts of a quality assurance program.
Independent checks of the data and
calculations are recommended to
ensure that the reported data are both
accurate and precise.

7.  Maintenance Section 2.2.7
recommends periodic  maintenance
schedules to ensure that the
equipment  is capable  of performing as
specified.

8.  Assessment of Data for Accuracy
and Precision Sections 2.2.8 and
2.2.9 describe the  assessment of the
data for accuracy and precision,
respectively. Independent audit
activities provide accuracy checks of
flow rate measurements, filter
weighings,  and data processing.
The precision check is performed by
using collocated samplers. The
expected agreement between two
collocated samplers is ±15%.

9.  Reference Information Section
2.2.10 discusses the traceability of
measurements to established
standards of higher accuracy, a
necessary prerequisite for obtaining
accurate data.
  Sections  2.2.11 and 2.2.1 2 contain
the Reference Method and pertinent
references.
  Section 2.2.13 provides blank data
forms for the convenience of the user.

-------
                              Jan.1983
                                                                              Section 2.2.1
                          1.0   Procurement of Equipment and  Supplies
    Specifications for equipment and
  supplies for monitoring ambient air for
  total'suspended particulates (TSP) are
  provided in the Reference Method, as
  reproduced in Section 2.2.11.
    Upon receipt of the sampling
  equipment and supplies, appropriate
  procurement checks should be
  conducted to determine their
  acceptability, and their acceptability
  or rejection should be recorded in a
  procurement log. Figure 1.1  is an
  example of such a log,  and Section
  2.2.13 provides a blank copy for the
  Handbook user. This log will serve as
  a permanent record for future
  procurements and for any  fiscal
  projections for future programs. It will
  also help to provide continuity of
  equipment and supplies. Table 1-1
  provides a matrix of the activities
  involved in the procurement of
  equipment and supplies.
          The following list of equipment,
        apparatus, and supplies provides a
        reference to sections and subsections
        within this part of the Handbook to
        guide the user to specific checkout
        procedures. Here and throughout the
        balance of the text,  "section" refers to
        the primary divisions of Section 2.2;
        "subsection" refers to the
        subdivisions within  these sections.
                    Item
                                           Section
Analytical balance
Relative humidity indicator
Elapsed-time meter
Timer
Flow rate transfer standard
Sampler
Filter
Sampler motor
Faceplate gasket
Rotameter
Sampling head
Motor gasket
Flow transducer and recorder
2.2.2
2.2.2
2.2.2
2.2.2
2.2.2
2.2.2
2.2.3
2.2.7
2.2.7
2.2.7
2.2.7
2.2.7
2.2.7
2.1
2.2
2.3
2.4
2.5
2.6
3.1, 3.3
7.1
7.2
7.3
7.4
7.5
'7.6
 Table 1.1  A ctivity Matrix for Procurement of Equipment and Supplies
          Equipment
Analytical balance
Elapsed-time meter
 Timer

 Orifice calibration unit (flow
 transfer standard)
Sampler
Relative humidity indicator
                                Acceptance limits
 Indicated weight =
 standard weight rO.OOOS
 g for three to five standard
 weights over sample filter
 weight range
 24 h ~2 min
24 h x3Q min

Calibration flow rate —
actual flow rate ±2%
                               Sampler complete; no
                               evidence of damage: flow
                               -1.1-1.7 m3/min
Indicator reading -
psychrometer reading ±Q%
                            Frequency and method
                            of measurement
                            Action if
                            requirements
                            are not met
On receipt, check against
weights of known
accuracy.


On receipt, check against
standard timepiece of
known accuracy.
On receipt, check against
elapsed-time meter.
On receipt, check against
flow-rate primary standard.


On receipt, observe visually
and check operation of all
components.
                                                           On receipt, compare with
                                                           reading of a wet bulb/dry
                                                           bulb psvchrometer.
                                                                                       Request recalibration
                                                                                       by manufacturer/supplier.
                                                                                       Adjust or reject.
Adjust or reject.

Adopt new calibration
curve if no evidence of
damage: reject if damage
is evident.
Reject or repair.
                           Adjust or replace to attain
                           acceptance limits.

-------
                       Section 2.2.1
                                                                        Jan. 1983

Item description
Hl-VOLSWHSl

Quantity
/z
/oo
Purchase
order
number
JJSJL

Vendor
' ~w*

Date
Ordered
&-/-75
6-/-T5
Received
(o ~IO~ 75

Cost
^/

Dispo-
sition
Ace.
Ace.
Comments

•
Figure 1.1.     Example of a Procurement Log.

-------
                              Jan. 1983
                                                                             Section 2.2.2
                                    2.0   Calibration  of Equipment
    Before a TSP sampling program is
  undertaken, a wide variety of
  sampling and analysis equipment
  must be calibrated. The calibration
  activities are summarized in Table 2.2
  at the end  of this section. Many of
  these activities will also serve as
  initial acceptance checks. All data
  and calculations required for these
  activities should be recorded in a
  calibration  log book in which a
  separate section is designated for
   each apparatus and sampler used in
   the program.


   2.1   Analytical Balance
     The calibration should be verified
   (1) when the analytical balance is first
   purchased, (2) any time it has been
   moved or subjected to rough handling,
   and (3) during routine operations
   when a standard weight cannot be
   weighed within ±0.5 mg of its stated

High-Volume Filter-Weighing Quality Control Log
                     C/ass-S weights, g
wefght. A set of three to five standard
weights covering the range normally
encountered in weighing filters should
be weighed. If the weighed value of
one or more of the standard weights
does not  agree within —0.5 mg of the
stated .value, the balance should be
recalibrated or adjusted by the
manufacturer. The  results of all
balance checks should be recorded in
a log book such as  the one shown in
Figure  2.1.
Date
7/29/74
7/29/74
7/29/74
7/30/74
7/31/74
7/31/74
7/31/74
7/31/74
7/31/74
8/1/74
8/1/74 •
8/1/74
8/1/74 '
8/1/74
8/1/74 '
8/1/74
8/2/74
8/2/74
8/2/74
8/2/74
8/2/74
8/2/74
8/2/74
8/3/74
8/3/74
8/5/74
8/5/74
8/5/74
8/5/74
8/5/74
3/5/74
8/5/74
8/6/74
8/6/74
8/6/74
8/6/74
8/6/74
8/6/74
8/7/74
8/7/74
8/7/74
8/7/74
8/7/74
8/8/74
9/24/74
9/26/74

Time
1 1:07
12:08
2:40
4:03
9:57
10:56
1 1:57
2:04
3:05
9:03
10:05
11:10
12:12 :
7:43
2:42
' 3:45
8:54
9:56
10:59
12:16
1:55
3:03
4:00
8:41 •
11:15
8:42
9:45
10:44
1 1:46
1:16
2:21
3:15
9:37
1 1:05
12:10
2:10
3:09
4:05
8:50
9:46
1:10
2:20
3:25
9:46
3:50
3'OJ

0.5000
0.5000
0.5002
0.5000 .
0.4996
0.4997
0.4995
0.4996
0.5001
'0.5000
0.4998
0.4999
0.5000
0.4998 '
O.SOOO
0.5001
0.5001
0.5000
O.SOOO
0.5003
0.5001
0.4999
0.5000
O.SOOO
0.4999
0.5002
0.5001
O.SOOO
0.5000
0.5000
0.5OO1
0.5001
O.SOOO
0.4999
0.5000
0.4999
0.5000
O.SOOO
O.SOOO
0.5000
0.4996
0.5001
0.5001
0.5002
O.SOOO
0.5OO1
	 : 	 _
1.0000
1.0002
1.0003
1.0000
09999
1.0000
. 0.9996
0.9998
1.0000
1.0000
0.9997
0.9997
1.0001
0.9997
1.OO01
1.0001
J.OOOO
1.0001
1.0000
0.9999
1.0002
. 1.0002
0.9999
0 9998
0.9996
1.0002
1.0000
1.0000
1.0000
1.00O1
1.0000
1.0000
1.0000
1.00OO
0.9998
0.9998
0.9998
J.OOOO
1.0000
1.0002
O.9998
1.0OOO
J.OOOO
1.0001
J.OOOO
	 1.00O1 	
	 1-°°01 	
2.0000
2:0000
2.0001
1 9999
2.0002
2.0000
1.9996
1.9998
2.0002
2 0000
1.9997
1.9997
2.0001
1.9998
2.0002
2.0001
2 000 1
2.0001
2,0001
1.9998
2.0002
2.0001
2.0O01
1 9999
1.9998
2 0002
2.0000
• 2.0000
2.0000
2.0000
2.0000
2.0001
2 0001
2.00OO
1.9997
1.9999
2.OOOO
2.0000
2 0000
2.0003
1.9996
2.0000
2.0000
2 0000
2 0000
2.0001
2. 000 1
Technician
BSM .
DEK
DEK
JLK
DEK
DEK
DEK
DEK
DE'K
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK
DEK



Figure 2.1.    Example of balance performance record.

-------
                         Section 2.2.2
                                                                           Jan. 1983
 2.2  Relative Humidity
 Indicator"
   The relative humidity indicator used
 for monitoring the filter conditioning
 environment should be checked
 against a wet bulb/dry psychrometer
 or the equivalent every 6 months. At
 least a two-point calibration should be
 made by comparing readings made in
 the conditioning environment against
 those made outdoors or perhaps just
 outside of the conditioning room. If
 the difference between the indicator
 and the corresponding psychrometer
 readings is within ±6%. it Is  all right
 to continue using the relative
 humidity indicator; if not, the indicator
 must be calibrated or a new one must
 be purchased. Record the results of
 (he relative humidity indicator checks
 in the calibration log book.

 2.3   Elapsed-Time Meter
   Every 6 months the elapsed-time
 meter should be checked against a
 timepiece of known accuracy, either
 on site or in the laboratory. If the
 indicator shows any signs of being
 temperature-sensitive,  it should be
 checked on site during each season of •
 the year. A gain or loss >2 min/24-h
 period warrants adjustment or
 replacement of the indicator. The
 results of  these checks should be
 recorded in the calibration log book.

 2.4   On-Off Timer
  The on-off timer should be
 calibrated and adjusted quarterly by
 using a calibrated elapsed-time meter
 as the reference. An example
calibration procedure for one type is
presented below. Figure 2.2 depicts
 the connection diagram for calibration
of a particular kind of timer. The steps
in the procedure are:
  1,  Plug a correctly wired timer into
     an electrical outlet.
  2.  Set the timer to the correct time.
  3.  Set the on and off time-trippers
     for a 24-h test.
  4.  Plug the test light  into one of the
    output plugs, and plug an •
    elapsed-time meter into the
    other,
  5, Check the system by manually
    turning the switch on and off.
  6, Allow the system to operate for
    the 24-h test period, and
    determine the time elapsed on
    the elapsed-time meter. If the
    elapsed time is 24 h ±30  min, the
    timer is acceptable for field use;
    if not, adjust the tripper switches
    and repeat the test. Record the
    calibration data in a timer
    calibration log such as that
    shown in Figure 2.3. Section
  Indicator Lamp
  On-Off Timer
 l± 15 min/24 h)
                                                           Elapsed- Time Meter
                                                             (±2 min/24 h)
 Figure 2.2.  Diagram of a timer calibration system.
     2.2.13 provides a blank copy for
     the Handbook user.

 2.5  Flow Rate Transfer
 Standard
   Calibration of the high-volume
 sampler's flow indicating or control
 device is necessary to establish
 traceability of the field measurement
 to a primary standard via a flow-rate
 transfer standard. The calibration
 procedure provided here applies to a
 conventional orifice-type flow transfer
 standard. Other types of transfer
 standards may be used if the
 manufacturer or user provides an
 appropriately modified calibration
 procedure that has been approved by
 EPA (see 40 CFR, Part 58, Appendix
 C, Section 2.8).
   Upon receipt and at  1 -year
 intervals, the calibration of the
 transfer standard orifices should be
 certified with a positive displacement'
 standard volume meter (such as a
 Rootsmeter) traceable to the National
 Bureau of Standards (NBS). Orifice
 units should be  visually inspected for
 signs of damage before each use, and
 they should be recalibrated if the
 inspection reveals any nicks or dents
 in the orifice.

  The following  equipment is required
for certification of an orifice transfer
standard.

  1.  Positive-displacement, standard
volume  meter (such as Rootsmeter)
traceable to NBS.
  2.  High-volume air pump (high-
volume sampler  blower).
  3.  Resistance  plates  or variable
voltage regulator.
  4.  Stopwatch
  5.  Thermometer
  6.  Barometer
  7.  Manometers [1 mercury (Hg), 1
water, or equivalent).
   The following step-by-step
 procedure for certification of an orifice
 transfer standard is adapted from the
 Reference Method.1 An orifice.
 transfer standard certification
 worksheet (Figure 2.4) is  provided for
 documentation of certification data.

   1.  Record on the certification
 worksheet the standard volume  meter
 serial number; transfer standard type,
 model,  and serial  number; the person
 performing  the certification; and the
 date.

   2.- Observe the barometric pressure
 and record it as PI (item 8)..
   3. Read the .ambient temperature in
 the vicinity of the  standard volume
 meter and record it as T, (item 9) (K =
 °C + 273).
   4. Connect the orifice transfer
 standard to the inlet of the standard
 volume meter. Connect the mercury
 manometer to measure the pressure
 at the inlet of the  standard volume
 meter. Connect the orifice (water)
 manometer to the pressure tap on  the
 orifice transfer standard. Connect a
 high-volume air pump (such as a
 high-volume sampler blower) to  the
 outlet side of the standard volume
 meter. (See Figure 2.5 for an example
 of the calibration setup.)

  5.  Check for leaks by temporarily
clamping both manometer lines (to
avoid fluid loss) and blocking  the
orifice with a large-diameter rubber
stopper, wide cellophane tape, or
other  suitable means. Start the high-
volume  air pump and note any change
in the standard volume meter reading.
The reading should remain constant.
If the  reading changes, locate any
leaks by listening  for a whistling
sound and/or retightening all
connections, making sure that all
gaskets  are properly installed.

-------
                  Jan. 1983
                                                                     Section 2.2.2
    •§ 2
    •i 1
 Ja-S
 •Q 
-------
                       Section 2.2.2
                                                                         Jan. 1983

I

•2
v\
T
<

rx


Wl
^.

en



-

oil (C
-x _! 1
a:
'==0>'
Pressure
drop across
orifice
AW
\Atinl or D (cm)
of water
2 1
2s*
S ads

|U| 1
iK-i
-^ s
it - *
"*
V. =>
2-S §• -^
!j**fi
Ifl^S
o/\j uny
c

,8
IX

Q
<0
O
O

5

8
ai
CO
01
8
3
1
~-
£

S
„.

Q
s?
Q
ti

§

§


-------
                               Jan. 1983
                                                                               Section 2.2.2
   Mercury
  Manometer






ft p




\ '<
"il
jj
lilj
g?;;
zn ,- >
n
                                                     Thermometer
                       Barometer
                                                  Orifice Transfer Standard

                                                         Positive
                                                         Displacement
                     Dial No. 5
                     Uncompensated
                                                            Variable
                                                            Voltage
                                                           Transformer
 Figure 2.5.  Example of orifice transfer standard calibration setup.
  6. Check the level of the positive -
displacement meter table, and adjust
the legs if necessary.
  7. After satisfactorily completing
the leak check, shut off motor,
unclamp both  manometer lines, and
zero the water and mercury
manometers by sliding their scales
until the zero is even with the
meniscus, as illustrated in Figure 2.6.
  8. Achieve the appropriate flow rate
through the system, either by means
of the variable flow resistance in  the
transfer standard or by varying the
voltage to the  air pump. (Use of
resistance plates  is discouraged
because the leak  check must be
repeated each time a new resistance
plate is installed.) At  least five evenly
distributed different but constant flow
rates are required, at least three of
which must be in the specified flow
rate interval (1.1 to 1.7 mVmin [39-
60 ftVmin]).
   9. Start the blower motor, adjust
 the flow,  and allow the system to run
 for at least 1 min to attain a constant
 motor speed. Observe the standard
 volume meter dial reading and
 simultaneously start the stopwatch.
 Error in reading the  meter dial can be
 minimized by starting and stopping
 the stopwatch on whole numbers
 (e.g., 0046.00).  .

   10. Record the initial meter reading
 (V,) in Column 1. Maintain this
 constant flow rate until at least 3 m3
 of air have passed through the
 standard volume meter. Record the
 standard volume meter inlet pressure
 manometer reading as AP (Column 5),
 and the orifice manometer reading as
 AH (Column 7).  Be sure to indicate
 the correct units of measurement.
  1 1. After at least 3 m3 of air have
passed through the system, note the
standard volume meter reading and
  simultaneously stop the stopwatch.
  Record the final meter reading (Vi) in
  Column 2 and the elapsed time (t) in
  Column 3.
    12.  Calculate the volume measured
  by the primary standard volume meter
  (Vm) at meter conditions of
  temperature and pressure (using
  Equation 1 of the work sheet) and
  record in Column 4.
               Vm = V, - V,
                          Equation 2-1
    13.  Correct this volume to standard
  volume (std m3) by  using  Equation 2
  of the work sheet:
                                                                                     Vsta = V,
                                                      , /PI . APV TsjA
                                                       V  p- A  T, /
                          Equation 2-2
  where:
    Vs,d = standard volume, std m3;
     Vm = actual volume measured by
         the primary standard volume
         meter, m3 (Column 4 of work
         sheet)
     Pi = barometric pressure during
         calibration, mm (in.) Hg (Item
         8 of work sheet)
    AP=differential pressure at inlet to
         primary standard volume meter,
         mm (in.) Hg (Column 5 of work
         sheet)
    Psia = 760 mm Hg (29.92 in. Hg)
 .   TSM = 298 K
     Ti = ambient temperature during
         calibration, K (Item 9 o.f work
         sheet).
   14. Calculate the standard
 volumetric flow rate (std mVmin) by
 using Equation 3 of the work sheet:

              CUid = YJU<<
                    t    Equation 2-3
 where:
    QSW = standard volumetric flow
         rate, std mVmin  at 760  mm
         Hg and 298 K
       t = elapsed time, minutes
   15.  Record Qstd to the nearest 0.01
 std m3/min in column  6 of the work
 sheet. Repeat steps 9 through 15  for
 at least four additional constant flow
 rates evenly spaced over the
 approximate range of 1.0 to 1.8 std
 mVmin  (35-64 ftVmin).
   16.  For each flow, compute
 \/AH (P,/P,,d) (298/T,)  (Column 7a),
 and plot these values against Qstd  as
 shown in Figure 2.7. Be sure to use
 consistent  units (mm or in. Hg) for
 barometric pressure. Draw the orifice
 transfer standard certification curve or
 calculate the linear least squares
slope (m) and intercept (b) of the
certification curve:

 \/AH (P,/P3,d) (298/T,) = m Q,,d + b.
 A  certification graph should be
 readable to 0.02 std mVmin.

-------
                          Section 2.2.2
                                                                             Jan. 1983
           •3-r


           • 2-\


           • H

            O-i

           • 7
   - 2 —
   - 1 -
                                   2-E
        Mercury
      Manometer
        Zeroed ,__
                           , = 70mm
  Water
Manometer
  Zeroed
 Figura 2.6. How to read mercury and water manometers.
         40
         3,0
         2.0
         10
                                                                         -4-
    •3

     2 •

    • ; •

     o


   =- t -


   r 2~

   =- 3-

     4-
                                                                                                -3-
   - 2-
^L-J
 Mercury
Manometer
 Reading
Pm = 7Omm
                                                                 =• 0-
                                                                 E-2-E
                                                                                               - 3 -
                                                                              />, = 3 //7.
  Water
Manometer
 Reading
P< = 3.0in.
          0
           o.o
                                                      Slope (m) =  2.062
                                                    Intercept fb) = -0.056
             I I  I I I i i  i  l M i  i i  I i  i  i i  I i  i  i i  I i  i t  i  I i  i i  i I
                                         _1_L
                                                   17.  If any calibration point does not
                                                fall within ±2% of the line, rerun that
                                                point,  recalculate, arid replot. The
                                                percent deviation can be calculated by
                                                comparing each Y from Column 7a
                                                against the corresponding Ycai
                                                calculated from the slope and
                                                intercept  using Equation 2-4:
                                                                     std + b
                                                                         Equation 2-4
                                                The percent deviation 'for each point is
                                                then calculated using Equation 2-5.

                                                       % deviation =Y - Ycai Y 1 00
                                                                    Yca,
                                                                         Equation 2-5
                                                  18.  For subsequent use- of the
                                                transfer standard, calculate Qsia as
                                                          m        x.	f
                                                                         Equation 2-6
                                                or determine Q5td for each value of:

                                                                          J\
                   0,25    O.50    Q.75    1.00    1.25    1.50
                                                                  1.75
Figure 2.7,  Example of on/ice transfer standard calibration relationship.
                         JWV  T2 ,
         from the certification graph.
         where:
            P2 = barometric pressure at time of
    2.00       Hi-Vol calibration
            TZ = temperature at time of Hi-Vol
               calibration

-------
                              Jan. 1983
                                       Section 2.2.2
  2.6   Calibration of High-
  Volume Sampler
    Each high-volume sampler must
  incorporate a flow rate measurement
  device capable of indicating the total
  sampler flow rate. This device may be
  an electronic mass flowmeter, an
  orifice or orifices located in the
  sample air stream together with a
  suitable pressure indicator (such as a
  manometer or an aneroid pressure
.  gauge), or any other type of flow
  indicator (including a rotameter)
  having comparable precision and
  accuracy. It must be possible to
  calibrate the flow rate measurement
  device to a flow rate that is readable
  (in corresponding units) to the nearest
  0.02 std mVmin. A pressure recorder
  with an orifice device that provides a
  continuous record of the flow may be
  used.
    The concentration of TSP in the
  ambient air is-computed as the mass
  of collected particles, divided by the
  volume of air sampled, corrected to
  standard conditions of 760 mm Hg
  and 298 K, and then expressed in
  micrograms per  standard cubic meter
  (fjg/sld m3). When samples are
  collected at  temperatures and
  pressures significantly different from
  standard conditions, the corrected
  concentrations may differ
  substantially from actual
  concentrations (micrograms per actual
  cubic meter), particularly at high
  elevations.
   Calibration of a high-volume
 sampler refers to calibration of the
 sampler's flow rate indicator so that it
 provides accurate measurements of
 the sample flow rate from which the
 volume of the sampled air can be
 calculated. Details of the calibration
 procedure vary somewhat depending
 on (1) the type of flow indicator used,
 (2) whether the sampler is equipped
 with an automatic flow controller, and
 (3) whether the calibration is to
 incorporate the geographical average
 barometric pressure and seasonal
 average temperature at the sampling
 site. The basic procedure for nonflow-
 controlled samplers is given in
 Subsection 2.6.2, whereas the
 variations in  the procedure necessary
 for flow-controlled samplers are
 presented in Subsection 2.6.3.
   Orifice-type flow indicators are
 sensitive to changes in both
 temperature and barometric pressure.
 Because ambient temperature and
 barometric pressure vary from day to
 day, the calibration procedure
 contains a formula to correct for this
 variability. Errors resulting from
 normal daily  fluctuation  are relatively
 small, however, compared with
 barometric differences due to
 elevation and seasonal temperature
 changes. Thus, if the modest errors
 due to daily changes are acceptable,
 the average barometric pressure for a
 given elevation and the seasonal
 average temperature for that location
 can be incorporated directly into the
 sampler calibration with little error
 being introduced in  the calculated
 flow rate.
   When this is done, the sampler is
 calibrated for the average temperature
 and pressure conditions at the  site,
 and no further temperature or
 pressure corrections are needed for
 the flow indicator reading to be used
 to determine the sampler flow  rate.
 The relationship between the flow
 indicator reading and the standard
 volume flow rate then becomes a very
 simple one. This relationship also can
 be easily reduced to a simple three-
 column table (indicator reading,
 winter flow rate,  and summer flow
 rate) suitable for  use even  by
 nontechnically oriented operators.
   The average barometric pressure for
 a site can  be estimated from the
 altitude of the site, either by using an
 altitude-pressure table or by reducing
 the sea level pressure of 760 mm Hg
 by 26 mm Hg for each 305 m (1000
 ft) of altitude. The average pressure
 could also be determined by averaging
 onsite barometer readings  or nearby
 weather station or airport
 measurements (station pressure,
 uncorrected) over several months. The
 seasonal average temperature for a
 site can be estimated from onsite
 temperature readings or nearby
 weather station records over the
 season. Ideally; the average
 temperature should reflect the
 temperature at the time of day  at
 which the  flow indicator would
 normally be read; however, an
 average determined  from 24-hour
 mean temperature records  would be
 acceptable. For most sites,  two
                                       seasonal average temperatures
                                       (summer and winter) are sufficient;
                                       for sites where climatic changes are
                                       severe, however, four seasonal
                                       average temperatures may be needed
                                       to accommodate the changes. Where
                                       computers are used to process TSP
                                       data, monthly average temperatures
                                       could be used. Ideally, the
                                       seasonal average temperature
                                       should generally be within ±15°C of
                                       the local ambient temperature at the
                                       time the flow indicator is read. If daily
                                       temperature changes at the site are
                                       too drastic to be represented by a
                                       seasonal average (±15°C) actual
                                       temperature corrections should be
                                       used each time a flow reading is-
                                       obtained.'
                                         Once a decision has been made on
                                       whether to incorporate an average
                                       barometric pressure and a seasonal
                                       average temperature into the
                                       calibration, the appropriate expression
                                       for plotting or calculating the sampler
                                       calibration can be selected from Table
                                       2-.1. The use of this  expression is
                                       explained in Subsection 2.6.2.

                                       2.6.1   Calibration Schedule - High-
                                       volume-sampler flow-rate devices
                                       should be calibrated with  a certified
                                       flow-rate transfer standard such as an
                                       orifice calibration unit (1)  upon
                                       receipt, (2) after motor maintenance,
                                       (3) any time the flow rate device is
                                       repaired or replaced, and (4) any time
                                       the difference between the sample
                                       flow rate and a one-point  audit
                                       deviates more than ±7 percent.

                                       2.6.2  Sampler Calibration Procedure
                                       - The procedures for multipoint
                                       calibration of a high-volume sampler
                                       are specified in 40 CFR 50, Appendix
                                       B (reproduced in  Section 2.2.11).  To
                                       facilitate these procedures, calculation
                                       data forms have been developed to aid
                                      -in making the calibrations. These
                                       forms also may be used for the
                                       calibration of other types of high-
                                       volume flow measuring devices.
Table 2.1.    Expressions for Plotting Sampler Calibration Curves

  Type of sampler     For actual pressure          For incorporation of
 flow rate measuring
	device
                        and temperature
                          corrections
Mass flowmeter

Orifice and pressure
  indicator

Rotameter, or orifice
  and pressure
  recorder having
  square root sca/e"
geographic average pressure and
 seasonal average temperature
                               I
                                                          I
                        IP2\(298
"This scale is recognizable by its nonuniform divisions: it is the most commonly
available for high-volume samplers.

-------
                                                        8
•*-*-•

     the
     the
                                                                             Jan. 1983
     '9h-vo/ume s/tesm  f ard for
   ca''bration priced* Per' and
   ""Portance The Ja,^ 3re of P
   "fleeted by the jnSru'S °f the
   dependent on th ns*ru
  i_  —   '"JW I'f^to i *j~       '^OvJ Fr]
 the flow recorder and r3"0" f/; fr°m
 column 4 Tar, .1     record unrfer

 aSL^ befo?taktgf'e^c;ercoro^


and
                                                                        the
                                       9raphica(/y
                                       Certification
       res 2.8n      h5 Sown
    he cottecifon and 1.  9'V6n to a'"d  '
 a'«8r to the
            uatavaPo/nt.
        aPpropr/ate Ln    fr°m the




        issure and seaSna? bar°me
          perature are tn h  • avera9e
           the          ° bel
                                                                                                  3nd a
                                       e<  nd
                               expressions to
                                                                   '-'ate

                                                                              with  f     " SamP'ers
                                                                           may be ca SW C0ntro"''"9
                                                                           of      '  brated eith
                                                                  ca'/bration, th^ ?nCiCator- After
                                                                  m-echan/s         -
                                                                               ,

-------
                                 Jan. 1983
                                                                                  Section 2.2.2
I
         --
oo
.<,
^
t»

I.
  ui
C    1
I    I
                  00

       CO
       Oo
            81
              CO

             -Q
   <0
         91


         I

         co
                                S!

                                S3
Hg
760
                                     •a
                                     c
                                       '§

                                       1
                                        =».
                           to  5  .
                           S  *-
                           4» w CN
                           *~ O
                           ?> 2 4>
                             0) -Q
                                     U)
                                     V)
pe
                           .^   a
                           S   o>
                           •U   Cn
                           s   ?
                           o
                            10
                           -Q
                            QJ

                            S1
                                    c
                                    o
Seasonal ave
                                                         a
                                                                      a
                                                  a
K ,
-------
                         Section 2.2.2
                                                      10
                                                                            Jan. 1383
     6O.O
     50.0
     4O.O
    30.0
    20,0
     70,0
                                                      Slope (m) = 32.773
                                                      Intercept =   3.3'51
                                                     Corr. Coff. =    .9991

                                        i  I I  f  i i  I l  i  i i I i  ill Mi  i
        0,0     0.25    0.50    0.75     7.00     7.25     /.50     7.75    2.00
 Flgura 2.9.  Example of a high-volume sampler calibration relationship.
 provided the previous operating
 history of the sampler demonstrates
 that the flow rate is stable and
 reliable, In this case, the flow
 indicator may remain uncalibrated,
 but it should be used to indicate any
 relative change between initial and
 final flows, and the sampler should be
 recalibrated more often to minimize
 potential loss of samples because of
 controller malfunction. The following
 procedures should be used.
   1, Set the flow controller for a flow
 near the lower limit of the flow range
 (1.1 std mVmin) to allow maximum
 control range.
   2, Install a clean filter in the
 sampler and carry out steps 2 through
 5  and 7 through 9 of Subsection 2.6.2.
 No resistance plate should be used
 with the flow rate transfer standard.
   3, Following calibration, add one or
 two additional clean filters to the
 sampler, reconnect the transfer
 standard, and operate the sampler  to
 verify that the controller maintains the
 same calibrated flow rate; this is
 particularly important at high
 altitudes, where the flow control
 range may be reduced.
following procedures may be used.
(Refer to Figure 2.11, a photographic
copy of the rotameter, to identify the
working components in this
procedural step for adjusting the
rotameter.)
  1. Attach the rotameter to the high-
volume sampler motor.
  2. Turn on motor and adjust to
selected flow rate.
  3. If adjustment is necessary, hold
the rotameter vertically and loosen
the locking nut by turning it
counterclockwise.
  4. Turn the adjusting screw to the
desired setting (clockwise to lower the
ball, or counterclockwise to raise the
ball).
  5. Be sure the  ball continues to
read the desired setting after the
adjustment is made and as the locking
nut is tightened.
  6. Seal both the locking nut and the
adjustment screw with glue to as-sure
that the setting does not  change. Do
not cover the exhaust orifice.       . '
  7. Proceed with calibration of
rotameter as specified in  Subsection
2.6.2.
2.6.4  Rotameter Calibration
Procedure • High-volume samplers
equipped with rotameters are
calibrated by using the same
procedures and forms as specified in
Subsection 2,6 2. Should adjustment
of the rotameter be necessary, the

-------
                            Jan. 1983
                                                                                   Section 2.2.2
2. 7 0.
                              anrf or/y/ce- on/f assembled for 'calibration with flow

-------
                           Section 2.2.2
                                                           12
                                                                                   Jan. 1983
                 Spring-Clip
                  Backing Plate
                  Spring-Clip Support
   Adjusting Screw


       Locking Nut


    Exhaust Orifice
                                       Tapered Plastic Tube ~
                                                    Scale
                  Inlet Port
             Ball





      Base Screw

Backing Plate -  V
Figure 2,11.    Example of high volume sampler rotameter.

-------
                              Jan. 1983
                          13
                                                                              Section 2.2.2
 Table 2.2.    Activity Matrix for Calibration of Equipment
 Equipment
Acceptance limits
 .Frequency and method
 of measurement
Action if
requirements
are not met
 Analytical balance
 Relative humidity indicator



 On-off timer


 Elapsed-time meter



 Flow-rate transfer standard
Sampler
Indicated weight = true
weight ±0.0005 g
Indicator reading =
psychrometer reading ±6%
±30 min/24 h
±2 min/24 h
                                Indicated flow rate (from
                                previous calibration) =
                                actual flow rate ±2%
                                Indicated flow rate = actual
                                individual calibration
                                points ±5% of linearity
 Gravimetvic test-weighing
 at purchase and during
 periodic calibration checks;
 use three to five standard
 weights covering normal
 range of filter weights.

 Compare with reading of
 wet bulb/dry bulb psychro-
 meter on receipt and at
 6-mo intervals.
 Check at purchase and
 quarterly with elapsed-
 time meter.
 Compare with a standard
 timepiece of known
 accuracy at receipt and at
. 6-mo intervals.
 Check at receipt and at 1 -yr
 intervals against positive-
 displacement standard
 volume meter; recalibrate
 or replace orifice unit if
 damage is evident.
 Calibrate  with certified
 transfer standard on
 receipt, after maintenance
 on sampler, and any time
 audit deviates more than
 4.70/1
                                                                                         Have balance repaired
                                                                                         and/or. recalibrated.
Adjust or replace to attain
acceptance limits.
                                                        Adjust or repair.
                                                        Adjust or replace time
                                                        indicator to attain accept-
                                                        'ance limits.

                                                        Adopt new calibration
                                                        curve.
                                                        Recalibrate.

-------

-------
                               Jan. 1983
                                                                               Section 2.2.3
                                 3.0  Filter Selection and  Preparation
    Suppliers of glass fiber filters for
  measurement of TSP have two grades
  of materials—the standard or
  traditional grade that has been in use
  for more than 20 years and a spectro-
  quality grade. Because the spectro-
  quality grade contains less organic
  and inorganic contaminants, it is
  recommended for use where
  additional chemical analyses are
  anticipated. A filter with low surface
.  alkalinity is preferred to avoid positive
  interferences from absorption of acid
  gases while sampling. Ideally, surface
  alkalinity should be between pH 6.5
  and 7.5; however, most commercially
  available glass fiber filters have a pH
  of >7.5. Filters having a pH of
  between 6 to 10 are acceptable. An
  activity matrix for filter selection and
  preparation is presented as Table 3.1
  at the end of this section.

  3.1   Selection
   Only filters having a collection
  efficiency of >99 percent for particles
  of 0.3-pm diameter (as measured by
  the OOP tesrASTM-D2986-71) are to
  be used. The manufacturer should be.
  required to furnish proof of the
  collection efficiency of a batch of new
  filters. The collection efficiency should
  be recorded in the procurement log.
  Figure 1.1 of Section 2.2.1.
   Each filter should be visually
  inspected using a light table. Loose
 fibers should be removed with a soft
 brush. Discard or return to the
 supplier the filters with pinholes and
 other defects-such as tears, creases,
 or lumps.

 3.2  Identification for  Filters
 Not Numbered by the
 Supplier
  A serial number should be assigned
 to each filter. The number should be
 stamped on two diagonally opposite
 corners—one stamp, on each  side of
 the filter.  Gentle pressure should be
 used in application to avoid damaging
 the filter.

 3.3   Equilibration
  Each-filter should be equilibrated in
 the conditioning environment for 24 h
 before weighing to minimize errors in
 the weight; longer periods of
 equilibration will  not affect accuracy.
The conditioning environment
temperature should be  between 15°
  and 30°C (59° to 86°F) and should
  not vary more than ±3°C (5°F); the
  relative humidity (RH) should be
  <50% and not vary more than ±5%. A
  convenient working RH is 40%.

  3.4   Weighing
    Clean filters are usually processed
  in lots—that is, several at one time.
  Clean filters must not be folded or
  creased prior to their weighing or use.
  Before the first filter is weighed, the
  balance should be checked by
  weighing a standard Class-S weight
  of between 3 and 5 g. Actual and
  measured weights, the date, and the
  operator's  initials should be recorded,
  as shown in  Figure 2.1.
  , If the actual and measured values
  differ by more than ±0.5 rng (0.0005
  g), the values should be reported to
  the supervisor before proceeding.  If
  the actual and measured values agree
  within ±0.5 mg, each filter should be
  weighed to the nearest milligram.
  Each filter should be weighed within
  30 seconds after removing it from the
  equilibration chamber, and the tare
  weight and the serial number of each
  filter should be recorded in the
  laboratory log (Figure 3.1). Section
  2.2.13 contains a blank copy of Figure
  3.1  for the  Handbook user. Note:
 Silicone-treated high  volume filters
  have been found to have a static
 charge problem. This problem can  be
 eliminated by placing an antistatic
 device containing a low-level alpha
 radiation source within the balance
 chamber. These devices are
 commercially available.

 3.5   Handling
  A quantity of filters sufficient for a
 >3-mo period for each sampler should
 be numbered and weighed at one
 time. Pack the filters in their original
 container (or a box of  similar size) so
 that each filter is separated by a sheet
 of SVz-by-l 1 -in. tracing paper. Be sure
 the filters are stacked in the box in
 numerical order so that  the operator
 will use the proper filter first.
  In addition to the filters, the field
 operator  should be supplied with
 preaddressed return envelopes to
 protect the filters during mailing;
 these can be printed front and back to
 serve as a sample record data form,
as shown in Figure 3.2. Section
2.2.13 contains a blank copy of Figure
3.2 for  the Handbook user.

-------
                   Section 2.2.3
                                                                   Jan. 1983
aie
        5
        *••.

        I
        3
        §•
        t/5
        "5
e Para
Code
on
Fu
Laborato
Filter Preparation
must be free of p
  §"•§
  II
si*
« §"«•
>*•£
w Ql _ O
to 4; Q lo
2 »< 51 V
fA Humidity i. 5% A Temp
Balance must be checked
weighing (STD. WGT. 3
tO.0005 gl
Q
tu
I1

*
71
S
V^

~xh*
o^

fv
fr}
^

^
Cl<
>^.
V

^
^%
^
^*V




.
b












u











-































































•














































































































,.
.:'






































































































	 1















































































































































































u, .
•:



























	 ^























































.:,



,;























r





























""
























" •
.


























"•


	

























„, ,,
11
	



„;;,

























'

	 :





















'(,






















»'




„.
	


























11
„, 	
;,,..,
Hi.,,'.,..,

.
























., if.
j;!1!,:"
c
,:.O
2
3"
C;
,§
91
s
li
Q
5
C
91
cn
g
^
"^
fr
o
u:
                                                                                               ri	;,i|}
                         I
                             0-
                                                                                                I
                                                                                               2
                                                                                               o.

Q>


M i
3! ^

*• 
-------
                                  Jan. 1983
                                                                                       Section 2.2.3
   Comments
Figure 3.2.    Hi- Vol field data form.
                                                                                        Hi-Vol Data Record
                                                                    Project
                                                                    Station
                                                                    Site and/or Sampler No.
                                                                                       SAROAD Site Code
                                                                    Sample Date

                                                                    Filter No. 	
                                                                    Flow Reading initial
                                                                                final  _
                                                                    Average Flow Rate  .
                                                                   Running Time Meter initial
                                                                                       final _
                                                                   Total Sampler Time

                                                                   Total'Air Volume 	

                                                                   Net TSP Weight 	
                                                                   TSP Concentration
                                                                   Optional
                                                                                    Temperature
                                                                          initial .
                                                                         finaJ
                                                                         average
                                                                  Operator
    . std m3
. fjg/stcf mj
                                                                                                           Barometric Pressure

-------
                         Section 2.2.3
                                            Jan. 1983
 TabtaS. 1.    Activity Matrix for Filter Selection and Preparation
Activity
 Acceptance limits
 Frequency or method
 of measurement
Action if
requirements
are not met
Selection and collection
efficiency

Integrity

Identification


Equilbration
Weighing procedure
Handling
 Efficiency of>99% in 0.3-
 pm diameter particle
 collection.
 Nopinholes, tears, creases.
 etc.
 Identification number in
 accordance with specifica-
 tions
 Equilibration in controlled
 environment for >24 h;
 constant humidity chamber
 With FtH of <50%
 constant within ±5%;
 temperature between 15°
 and 3O°C with less than
 ±3°C variation
 Indicated filter weight
 determined to nearest mg
 within 30 s after removing
'from the equilibration
 chamber.
 Filter in protective folder:
 envelopes undamaged.
Manufacturer's proof of
OOP testASTM-D2986-71

Visual check of each filter
with tight table
Visual check of each filter
 The room or chamber
 conditions and the equili-
 bration period a re observed
 for each sample.
Observation of weighing
procedure.
Visual check of each filter.
Reject shipment or return
to supplier.

Discard filter.

Identify properly or discard
filter.

Repeat equilibration.
                                                                                         Reweigh after re-equilibra-
                                                                                         tion.
                                                                                        Replace undamaged filters.
                                                                                        discard damaged filters

-------
                              Jan. 1983
                                       Section 2.2.4
                                        4.0   Sampling  Procedure
    The activity matrix presented as
  Table 4.2 at the end of this section
  summarizes the sample collection
  activities and the operational checks.

  4.1   Filter Installation
    Care must be taken  to assure that
  the clean weighed filters are not
  damaged or soiled prior to installation
  in the high-volume sampler. They
  should be kept in a protective folder or
  box and must not be bent or folded.
  The use of filter cassettes (Figure 4.1)
  that can be loaded and unloaded  in
  the laboratory may be  used to
  minimize damage to the filter.
  Damaged or soiled filters must be
  discarded.

   The following procedures are used
  to install a filter.
   1. Open the shelter and remove the
 faceplate of the sampler by loosening
 the four wingnuts and swinging the
 bolts outward.
   2. Wipe all dirt from the support
 screen and faceplate.
   3. Center the filter with the rough
 side up on the wire screen so that the
 gasket will form an "airtight seal on
 the outer edge (1 cm) of the filter
 when  the faceplate is in position.
 When  aligned correctly, the edges of
 the filter will be parallel both to the
 edges  of the  screen behind it and to
 the faceplate gasket above it. Poorly
 aligned filters show uneven white
 borders {Figure 4.2) around the filter.

  4. Tighten the four wingnuts just
enough to prevent leakage when the
filter is aligned and the faceplate is in
Figure 4.1.    High volume sampler filter cartridge assembly.
 place. Excessive tightening may cause
 the filter to stick or permanently
 damage the gasket.
   5. Close the shelter and run the
 sampler for at least 5 min to establish
 run-temperature conditions.
   6. Record the flow indicator reading
 and, if needed,  the barometric
 pressure (P3 maiai) and the ambient
 temperature (T3 imnai), then stop the
 sampler. Note: No onsite pressure or
 temperature measurements are
 necessary if the sampler flow
 indicator does not require pressure or
 temperature corrections (e.g., a mass
 flowmeter) or if average barometric
 pressure and seasonal average
 temperature for the site have been'
 incorporated into the  sampler
 calibration. For  individual  pressure
 and temperature corrections, the
 ambient pressure and temperature at
 the time of the flow indicator reading
 can be obtained by onsite
 measurements or from a nearby
 weather station. Barometric pressure'
 readings obtained from  airports must
 be station pressure, not corrected to
 sea  level, and may  need to be
 corrected for differences in elevation
 between the sampler  site and the
 airport. For samplers  having flow
 recorders but not constant flow
 controllers, the average temperature
 and pressure at the site during the
 sampling period should  be estimated
 from U.S. Weather Bureau or other
 available data.
   7. Determine  the flow rate from the
 sampler's calibration relationship
 (Subsection 4.4) to verify that it is
 operating within the acceptable range
 of 1.1 to 1.7 mVmin (39-60 ftVmin).
 If  not within this range,  use a
 different filter or adjust the sampler
 flow rate. Warning: Substantial flow
 adjustments may affect the calibration
 of orifice-type flow indicators and  may
 necessitate their recalibration.
 •  8.  Record the  sample  identification
 information (filter number, site
 location or identification number,
 sample date) and the initial flow rate
 (or flow indicator reading and
 temperature .and barometric pressure
 if needed) on the Hi-Vol  field data
form (Figure 4.3). See  Subsection 4.7
for proper documentation.
 ,9.  Set the timer to start  and stop
the sampler such that  the sampler
runs 24 hours, from midnight to
midnight local time.

-------
                         Section 2.2.4
                                                                           Jan. 1983
Figure 4.2,   Nonunifarm borders resulting from poorly aligned filters.
4.2   Retrieval of Exposed
Filter and Post-Sampling
Checks
  1  As soon as practical following
the sampling period, run the sampler
lor ai least 5 mm  to reestablish run-
temperature conditions.
  2. Record the flow indicator reading
and. if needed, the barometric
pressure (P3 i,n=i) and the ambient
temperature (T3 lmai).
  3. Stop the sampler, remove the
faceplate, and lift the exposed filter
from the supporting screen by
grasping it gently at the ends, not at
the'corners.
  4. Check the filter for signs of air
leakage. Leakage may result from a
  worn faceplate gasket (Figure 4.4) or
  from an improperly installed gasket. If
  signs of leakage are observed, void
  the sampler, determine the cause, and
  take corrective actions before starting
  another sampling period. A gasket
  generally deteriorates slowly; thus the
  operator can decide well in advance
  (by the  increased fuzziness of the
  sample outline) when to change the
  gasket before a total failure results.
   5. Visually inspect the gasket face
  to see if glass fibers from the filter are
  being left behind due to
  overtightening of the faceplate
  wingnuts and the consequent cutting
  of the filter along the gasket interface.
   6. Check the exposed filter for
  physical damage that may have
  occurred during or after sampling.
  Physical damage after sampling would
  not invalidate the sample if all pieces
 of the filter were put in the folder;
 however, sample losses due to
 leakages during the sampling period
 or losses of loose particulates after
 sampling (e.g., loss when folding the
 filter) would invalidate the sample, so
 mark such samples "void" before
 forwarding them to the laboratory.
   7.  Check the appearance of the
 particulates. Any changes from
 normal color, for example, may
 indicate new  emission sources or
 construction activity in the area. Note
 any change on the filter folder along
 with any obvious reasons for the
 change.
   8. Fold the filter lengthwise at the
 middle with the exposed side in; if the
 collected sample is not centered on
 the filter (i.e., the unexposed border is
 not uniform around the filter), fold so
 that only the deposit touches the
 deposit.  Results of an improperly
 folded filter  are illustrated in Figure
 4.5, where smudge marks from the
 deposit extend across the borders; this
 can reduce the value of the sample if
 the analyses for which the sample
 was collected need to be divided into
 equal portions.
   9. Place the filter in its numbered
 folder.
   10. Determine the final flow rate
 from the sampler's calibration
 relationship  (see Subsections 4.3 and
4.4) and  record it on the data record
along with other pertinent information
(see Figure 4.3).
   11. Remove the sampler's flow
recorder  chart and place the chart
inside the filter folder with the inked
side against  the folder and the
backside  against the filter.
4.3  Flow Readings
4.3.1  Rotameters -Jo obtain a valid
measurement, make flow rate

-------
                               Jan. 1983
                                                                               Section 2.2.4
                                                               Project

                                                               Station
                                                                                 Hi-Vol Data Record


                                                                        SPC.CIAL  3TUM
   Comments
      CITY
      CUEAN//V6   ON.
                                                              Site and/or(
                                                                                SAROADSiU Code
                                                           .   Sample Date


                                                              Filter No.
                                                              Flow Reading initial    '•

                                                                         final     /'
                                                              Average Flow Rate
                                                                                  /
                                                              Running Time Meter initial    OOOO

                                                                               final
                                                              Total Sampler Time
                       Total Air Volume,    /o69!>


                       Net TSP Weight 	Q.
                                                                                                             . std m3
                                                              TSP Concentration    /5

                                                              Optional
                                                                   . fjg/std m3
                                                                             Temperature
                                                                                                  Barometric Pressure
                                                                    initial

                                                                    final

                                                                  , average
                                                             Operator
 Figure 4.3.    Example of completed Hi-Vol field data form.
 measurements while the sampler is at
 normal operating temperature, after a
 warmup time of >5 min.
   1.  Connect the rotameter to the
 sampler with the same tubing  used
 during calibration, and place or hold
 it in a vertical position at eye level.
   2.  Read the widest part of the float
 (ball), and use the calibration
, relationship (see Subsection 4.4) to
 convert the reading to Qstd (mVmin)
 and record to the nearest 0.025ta
 mVmin.
  3. Measure the flow rates at the
beginning and end of each sampling
period. Observe the flow rate for >1
min after connecting the rotameter to
the sampler, before taking a reading. If
a gradual change in flow rate is
observed, do not take a reading until
equilibrium is reached; a gradual
change is usually observed when the
rotameter is at a substantially
different temperature from that of the
sampler exhaust air,  and thus
equilibration may require 2 or 3 min.
4.3.2  How Recorders - The
following procedure is for a high-
volume sampler equipped with a flow
recorder.
  2. Remove any moisture by wiping
the inside of the recorder case with a
clean cloth. Carefully  insert the new
chart into the recorder without
bending the pen arm beyond  its limits
of travel. An easy way to do this is to
raise the pen head by pushing in on
the very top of the pen arm with the
right hand while inserting the chart

-------
                         Section 2.2.4
                                                                            Jan. 1983
 Figura 4.4.    Example of air leakage around the filter due to worn faceplate gasket or to
              improper installation.
with the left hand Be careful not to
damage or weaken the center tab on
the chart, but be sure the.tab is
centered on the slotted drive so that
the chart will rotate the full 360
degrees in 24 h without binding or
slipping. A properly installed chart is
shown in Figure 4.6.
  3. Check to see that the pen  head
rests on zero (i.e., the smallest  circle
diameter on the chart). If not, tap the
recorder lightly to make certain that
the pen arm is free.
  4. Check the time indicated by the
pen. If it is in error, rotate the chart
clockwise by inserting a screwdriver
or coin into the slotted drive in  the  '
center of the chart face until the time
is correct. If the sampler is started
with a timer switch, the correct time
 is the starting time on the timer
 (usually midnight).
   5. Using an eyedropper, put a small
 amount of ink into the hole in back of
 the pen. tip. Use of cartridge-type pens
 will, minimize problems with inking.
   6. Turn the sampler on (never turn
 it  on until a filter is in place because
 the transducer  and recorder may be
 damaged), and  observe it long enough
 to know whether the transducer and
 recorder are operating properly.


4.4  Determination of Flow
Rates
  High-volume  sampler flow rate
readings must be converted to units of
std mVmin (25°C, 760 mm Hg) for
use in calculating TSP concentrations.
  Expressions for converting sampler
  flow rate readings (I) to standard
  conditions are .given in Table 4.1.
  Instructions for the use of this table
  and the flow measuring device
  calibration relationships (Figures 2.8
  or 2.9) to obtain the sampling flow
 . rate Qjtd (mVmin) are given in
  Subsections 4.4.1  and 4.4.2.
   No onsite pressure or temperature
  measurements are necessary if the
  sampler  flow indicator does not
  require pressure or temperature
  corrections (e.g., a mass flowmeter) of
  if average barometric pressure and
  seasonal average temperature for the
  site have been  incorporated into the
  sampler calibration. For individual
  pressure and temperature corrections,
  the ambient pressure and temperature
  at the time of flow indicator reading
  can be obtained by onsite
  measurements  or from a nearby
  weather station. Barometric pressure:
  readings obtained from airports must •
  be station pressure, not corrected to
 sea level, and may need to be
 corrected for  differences in elevation
 between  the sampler site and the
 airport. For samplers having flow
 recorders but not constant flow
 controllers, the  average temperature
 and pressure at the site during the
 sampling period should be established
 from Weather Bureag or other
 available data.

 4.4.1  , Samplers Without Continuous
 Flow Recorders - For a sampler
 without a continuous flow recorder,
 determine the appropriate expression
 to be used (from Table 4.1)
 corresponding to the one used in
 calibration (from Table 2.1). Using this
 appropriate expression, determine Qsta
 for the  initial flow rate from the
 sampler calibration curve, either
 graphically  or from the transposed
 regression equation (see Figure 2.8):
 Q5td=J_ ([Appropriate expression from
     m Table 4.1 ] - b)
                        Equation 4-1
 Similarly,  determine Qstd from the
 final flow reading,  and calculate the
 average flow.Qstd as one-half the sum
 of the initial and final flow rates.

 4.4.2  Samplers With Continuous
 Flow Recorders  - For a sampler with a
 continuous flow recorder, determine
 the average flow rate reading (I) for
 the period. Determine the appropriate
 expression from  Table 4.1
corresponding to the one used in
calibration (from Table 2.1). Then
 using this expression and the average
flow rate reading, determine Q3ld from
the sampler calibration relationship,
either graphically or from the

-------
                             Jan. 1983
                                                                             Section 2.2.4
Figure 4.5.    Example of smudged filter border resulting from an improperly folded filter.
Table 4.1.     Expressions for Determining Flow Rate During Sampler Operation

                      	    Expression	.__
                                                  For use when geographic.
                                                     average pressure
Type of sampler For actual pressure
flow rate measuring and temperature
device corrections
Mass flowmeter /
Orifice and pressure \l 1 ( P* \/25S\
indicator y [ p )\ j )
Rotameter. or orifice
and pre<;<:i/rf> 	 	
recorder having . /\l/P3\/298\
square root scale * y \ pM J\f3)
'This scale is recognizable by its nonuniform divisions
available for high-volume samplers.
temperature have been
incorporated into the
sampler calibration
1
1
and is the most commonly
  transposed regression equation (see
  Figure 2.8 and Equation 4-1 above):
    If the trace shows substantial flow
  change during the sampling period,
  greater accuracy may be achieved by
  dividing the sampling period into
  intervals, calculating an average
  reading for each interval, determining
•  Q5td for each interval, and finally
 •computing the average Qstd for the
  whole sampling period.
    Calculate the total air volume
  sampled by the following equation:
                V = Qstd t  Equation 4-2
  where: .
    V= total .air volume sampled, in •
       standard volume units (std m3.);
  Qstd = average standard flow rate, std
       mVmin;
     t = sampling time, min.

  4.5   Sampling Flow  Rate
  Checks
    The two types of sampling flow rate
  checks recommended are discussed in
  the following Subsections (4.5.1 and
  4.5.2).

  4.5.1   Initial Flow Rate Check - Initial
  flow rate measurements should be
  monitored for each sampler to
  determine whether corrective action  is
  needed.
    1. Record the initial and final flow
  rates for each sample in the log book
  maintained with the sampler. A
 sampler equipped with a continuous
 recorder should be observed for at
 least 5 min. before the initial flow rate
 is recorded.
    2.  Average the initial  flow rate
 measurements for the first four
 samples after each calibration. Check
 future initial flow rates that deviate
 more than ±10%  from this average
 for samplers on which a manometer
 or a flow recorder is used and ±15
 percent for samplers on which a
 rotameter is used. If the change has
 been gradual over time,  recalibrate. If
 large deviations occur between
 successive samples, repeat the flow
 reading after 5 minutes. If the second
 reading is within the above limits,
 continue normal operations;  if not,
 check the line voltage and/or replace
 the filter.
   3.  Perform a calibration check if
 neither of the above checks identifies
 the trouble. If the  calibration check is
 satisfactory, continue normal
 operations; if not,  perform a complete
 calibration.

 4.5.2  Operational Flow Rate Check -
 It is recommended that a one-point
 operational flow check be made on
 each sampler at least once every 2
 weeks. The purpose of this check is to

-------
                         Section 2.2.4
                                                                          Jan. 1983
                             •>  *   r^~ I  .   '  ™ , s"•.  r9
                              /-,/   ;-^..^ -^

                             ILT-U^M
Figure 4.6.   Flow rats recorder with chart installed.
track the in-control conditions of the
sampler calibration. The same flow
rate transfer standard used to
calibrate the high-volume sampler
may be used for the operational flow
check.
  1, Operate the sampler at its
normal flow rate with flow check
device  in place. Determine Q,,a' for the
check point from the  calibration of the
flow check device, and determine the
measured flow rate from the
sampler's calibration  (see Subsections
4 3, and 4.4) Use the following
procedure for plotting the check data.
  2. Calculate the percentage
difference (% D) between  the known
check flow measurement and the flow
measured by the sampler's normal
flow indicator (Equation 4-1). Let Qa
represent the known  flow rate and Qm
 the measured flow rat'e for the flow
 check:
             =/
               \
                        Equation 4-3
Thus if Qm = 1.48 mVmin and Qa =
1.42 mVmm
then:
    % D =/
i =/1.48 - 1.42\
  V   1.42  /
                      100 = +4%
If the % D is not within ±7 percent for
any one check, recalibrate before
resuming the sampling.
  3.  Record the Qm. the Qa, and the %
D on an X-and-R chart (Figure 4.7)
under "Measurement Result. Items 1
and 2." Record the % D in the cells
preceded by the "Range R." The % D
can be positive or negative, so retain
the sign of the difference, since it may
 indicate trends and/or consistent
 biases. More information on the
 construction of a quality control chart
 and the interpretation of the results
 are in Appendix H, Volume I of this
 Handbook.2
  4. Repeat the above for each
 operational flow rate check, plot all
 points on the chart, and connect  the
 points by drawing connecting lines.
 Tentative limits are ^4.7 percent
 (warning lines) and ±7 percent (out-
 of-control lines). Out-of-control points
 indicate possible problems in
 calibration or instrument errors. When
 out-of-control  results are obtained,
 recalibrate the sampler prior'to further
 sampling. After 15 to 20 points are
 plotted, new control and warning
 limits may be derived, as described in
 Appendix H of Volume I of this
 Handbook.2 Do not increase the
 control and warning limits; however,
 more stringent limits may be
 established.     _
  5. Forward the X-and-R  chart to the
 QA supervisor for review.

 4.6   Time Measurements
  Start and stop times for samplers
 not equipped with a timer switch  or
 an elapsed-time  meter are recorded
 by the operator who starts and stops
the sampler. If more than one
operator is involved, each should  set
 his/her watch to a common reference
to achieve accurate times;  such a
reference  could be an office clock that
is checked daily or the local telephone
company, which  gives the time of day.
The time measurement procedure is
as follows:

  1. Take the start and stop times for
 samplers equipped with timer
 switches from  the timers' start and
 stop settings.
  2. Check the timer clock, and set it,
 if necessary, for  the correct times at
 each filter change.
  3. Use an el-apsed-time meter'to
determine the  number of minutes
sampled because timers cannot be set
or read to within less than =30 min.

4.7   Documentation
  The following information should be
recorded on the filter folder or on a
field data record form (Figure 4.3) by
the persons indicated, and it should
be verified with a signature.

4.7.1   The Operator Who  Starts  the
Sample
  1.  Station location
  2.  Project number
  3.  Site number
  4.  Sampler ID  number
  5.  Filter number

-------
Jan. 1983
Section Z.Z.4

-------
                         Section 2.2.4
                                                                           Jan. 1983
   6, Sample date
   7, Initial flow reading (if using
 rotameter) and/or initial temperature
 and barometric pressure if required.
   8, Unusual conditions that may
 affect the results (e.g., subjective
 evaluation of pollution that day,
 construction activity, meteorology)  •
   9, Signature.

 4.7.2  The Operator Who Removes
 the Sample
   1. Elapsed time
   2. Final flow reading (or be sure
 that the flow rate chart accompanies
 the sample) and final temperature and
 barometric pressure if required.
   3, Existing conditions that may
 affect the results
   4, Signature.

 4.7.3  The Operator Who Transfers
 the Sample to the Laboratory Record
   1, Receiving date initialed
   2. Shipping date initialed.
 Table4.2.    Activity Matrix for Sampling Procedure
Activity
Filter installation
Flow checks
Elapsed time

Sample handling


Documentation
 Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
 Filter rough side up,
 centered on screen, edges
 parallel to edges of screen
 and to faceplate gasket;
 gasket tightened to prevent
 leakage
 1) Sampler flow rate within
 acceptable range of 1.1 to
 1.7 m3/min (39-60 ft3/min)
 21 Stabilized initial flow
 rate = established initial
• flow rate ±10% for
 pressure transducer or
 ±15% for rotameter
 3) Sampling time 24 ±1 h

 No evidence of malfunction
 in post-sampling check

 Names, sampling dates,
 times; sample, filter, and
, station numbers; unusual
 conditions: flow rates: and
 handling dates recorded on
 sample envelope	
                                                            Visually check each
                                                            exposed filter.
                                                            Check flow rate at each
                                                            filter change
Check on and off settings
of timers.
Visually check each sample
for tears, missing pieces.
or leakage.
Visually check each sample
data record
                            Void the filter: install
                            substitute filter.
                            1) Determine cause of flow
                            problem and correct;
                            measure line voltage,
                            change the filter, check
                            calibration and calibrate
                            sampler.
Reset timer.

Void the sample; correct
the cause of malfunction.

Complete or correct the
documentation; if unavail-
able, void the sample.

-------
                              Jan. 1983
                                                                             Section 2.2.5
                                       5.0   Analysis of Samples
   A matrix summarizing the major
  quality assurance activities for sample
  analyses is presented as Table 5.1 at
  the end of this section.


  5.1   Sample Documentation
  and Inspection
   Upon receipt of the sample from the
  field the following procedure should
  be followed:
   1. Remove the filter folder from its
  shipping envelope and examine the
  Hi-vol Field Data Record (Figure 4.3)
  to determine whether all data needed
  to verify the sample and to calculate
  the concentration have been provided.
  Void the sample if data are missing
  and unobtainable from the field
 operator or if  a sampler malfunction
 (e.g., faceplate gasket leakage)  is
 evident.
  .2.  Record the filter number on the
 Hi-vol Field Data Record and on the
 Laboratory Data Log {Figure 3.1).
   3.  Examine the shipping envelope.
 If sample material has been dislodged
 from the filter, recover as much as
 possible by'brushing it from the
 envelope onto the deposit on the filter
 with  a soft camel's-hair brush.
          4. Examine the filter. If insects are
        embedded in the sample deposit,
        remove them with Teflon-tipped
        tweezers, but disturb as little of the
        sample deposit as possible. If more
        than 10 insects are  observed, refer
        the sample to the supervisor for a
        decision to accept or reject  it.
          5. Record the data verification, the
        sample inspection, and removal of
        insects under "Remarks" in-the
        Laboratory Data Log.  •


        5.2   Filter Equilibration
          The following procedure should be
        used to equilibrate the exposed filters
        in a conditioning environment for 24
        h; up to 48 h may be needed for very
        damp filters.
          1. Use an  eqilibration chamber with
        a desiccant or an environmentally
        controlled weighing room to maintain
        an RH of <50 percent at 1 5° and
        30°C (59° to 86°F). An air-conditioned
        room may be used for equilibration if
        it can be maintained at an RH of
        <50% that is constant within =5%
        and an air temperature between 15~
        and 30°C that is constant within
        =3°C (5°F) while the filters are
 Table 5. J.    Activity Matrix for Analysis of Samples
                  equilibrating. A convenient working
                  RH is 40 percent. Keep a hygrometer
                  in the room.
                    2. Check the RH daily.
                    3. Record the hygrometer readings
                  and any equilibration chamber
                  malfunctions, discrepancies, or
                  maintenance in trie Laboratory Data
                  Log.

                  5.3   Gravimetric Analysis
                   A balance check should be
                  performed  as specified in Subsection
                  2.1.
                    1. Weigh the exposed filters to the
                  nearest milligram (mg)  on the
                  analytical balance.
                   2. Weigh the filters in the
                  conditioning environment if practical;
                  if not, be sure that the  analytical
                  balance is as close as possible to the
                  conditioning chamber where it is
                  relatively free of air currents and
                 where  it is  at or near the temperature
                 of the chamber. Weighing should take
                 place within 30 seconds after
                 removing filters from the equilibration
                 chamber.
                   3.  Record the weight in the
                 Laboratory Data Log and on the High
                 Volume Field Data Record.
A ctivity
                               Acceptance limits
                             Frequency and method
                             of measurement
                                                                                       Action if
                                                                                       requirements
Documentation verification
and sample inspection
Filter equilibration
Gravimetric analysis
" Complete documentation;
 no evidence of malfunction
 or sample loss; 24 h; RH <50% within
 ±5%; temperature constant
 within ±3°Cat 15°to30°C
 (59° to 86°F)
 Indicated weight obtained
 to nearest milligram within
 3O s after removal from
 equilibration chamber
Visually check all samples
and documentation.
                                                           For each sample, observe
                                                           room or chamber conditions
                                                           and equilibration period.
                                                           Observe filter weighing.
Void the sample
                            Repeat equilibration for 24
                            h at properly controlled
                            conditions.
                           Report to supervisor;
                           reweigh after equilibration
                           for 24 h at controlled

-------

-------
                            Jan. 1983
                                                                          Section 2.2.6
              6.0   Calculations  of TSP Concentrations and Data  Reporting
   A matrix summarizing the quality
 control activities for the calculations
 and the data-reporting requirements
 is presented in Table 6.1.

 6.1   TSP Concentration
   Equation 6-1 should be used to
 calculate the total air volume sampled.
              V = Q5td t  Equation 6-1
 where:
   V = Total air volume sampled, in
      standard volume units, std m3;
 Qstd = average  standard flow rate, std
      mVmin;
    t = sampling time, min.
   Equation 6-2 should be used to
 calculate the TSP sample
 concentration.
              = (W,-Wt)1Q8
                   V
                       Equation 6-2
       where
         TSP = concentration of TSP, //g/std
              m3,
          Wf=weight of exposed filter, g
          Wi = tare weight of filter, g.
       All original calculations should be
       recorded in the Laboratory. Data Log
       (Figure 3.1).

       6.2  Data Documentation
       and Reporting
         All daily concentration levels should
       be recorded in micrograms per
       standard cubic meter (//g/std m3),
       with .the required identifying
       information, on the SAROAD Daily
       Data form  (Figure 6.1). See AEROS
       User's. Manual, OAQPS No. 1.2-039,
       for detailed instructions.
Table 6.1.    Activity Matrix for Calculations and Data Reporting
Activity
                              Acceptance limits
                           Frequency and method
                           of measurement
                           Action if
                           requirements
                           are not met
Sample volume and   .
concentration .
Data documentation and
reporting
All needed data available

Complete documentation
for calculation of concen-
tration; all sample and data
identification numbers
matched; no evidence of
malfunction or sample loss;
all needed data available
Visually check data records
for each sample.
Visually check data record
and data log for each
sample.
Void the sample.

Void, the sample.

-------
                            Section 2.2.6
                                                                                  Jan. 1983
24'Hour or Greater Sampling Interval
t
Agency
S/rje/M'- Tokjn
City Name,
^jQ& /v&r"~fri SYV"CCv
Site Address
T5P 2.4 hour
Project
Day St
19 20 21
1
i

| 	


1
1
i
I 	
1
,

; .
0




\
O
1
i
i,
n

*


i

i







;





Time Interval
rsp
Name
PARAMETER
Cods
/
/
/
0
/
23 24 25 26 27

O
/
c
) /


O
Hf 28 29 . 3O 31 32
22 33 34 35 36.
O





O



O
O
1



-

t





0

O
1
i
1







a


,•





I

«








1














7




-------
                              Jan. 1983
                                        Section 2.2.7
                                            7.0   Maintenance
   Scheduled or preventive
  maintenance of the sampling
  equipment reduces voided samples,
  downtime, and remedial maintenance.
  Because the sampling equipment is
  operated only intermittently, the
  frequency of maintenance is a
  function of the actual hours of use.
  Normally, two or three preventive
  maintenance activities are required
  each year. When possible,
  maintenance is-best performed in the
  laboratory rather than in the field.
  Motors on which maintenance has
  been performed can then  be carried to
  the field for installation and
  calibration. Table 7.1 at the end of
  this section  summarizes the quality
  assurance activities of major
  maintenance checks. All maintenance
  activities should be recorded in a log
  book.
  7.1    Sampler Motor
   Motor brushes usually require
  replacement after 400 to 500 h of
 operation at normal line voltage (115
 V). The procedure is as follows:
   1. Replace the brushes before they
     are worn to the point that
     damage can occur to  the
     commutator of the Hi-Vol motor.
     The optimum replacement
     interval must be determined from
     experience.
   2. Follow the manufacturer's
     instructions for replacing the
     brushes.
   3. Recalibrate the high-volume
     sampler after the brushes are
     replaced.  Do not recalibrate the-
     motor until after an initial break-
     in period for the proper seating
     of the brushes against the
     armature; this period  usually
     requires running the sampler for
     several  hours against a
     resistance equivalent  to a clean
     filter or a  No. 18 calibration
     plate.
   4. Refer to the flow diagram in
     Figure 7.1 for the various steps
     required for motor maintenance.
   5. Record all sampler maintenance
     operations (with dates performed
     and the operator's initials) in the
     sampler log book and  on a
     gummed label  (Figure 7.2)
     attached to the sampler.
7.2  Faceplate Gasket
  A worn faceplate gasket is
characterized by a gradual  blending of
                            Open the Motor Housing
                                 Remove Motor
                              Inspect-Armature •
                        -*•  If Bad
                                                            -Replace Armature
                               Change Brushes

                                     I
                              ~"~ Check Motor -
     If Good
   Reassemble
    Final Test
      Field
   Calibration
                                                                If Bad
                                                              Remove
                                                             Usable Parts
                                                            Discard Motor
 Figure 7.1.  Flow diagram for high volume sampler motor maintenance.
the interface between the collected
particulates and the clean filter
border. Any decrease in the sharpness
of this interface indicates the need for
a new gasket.
  1. Remove the old  gasket with  a
knife.
  2. Clean the surface properly.
  3. Seal a new gasket to the
faceplate with rubber cement or
double-sided adhesive tape.
Hi-vol motor number.
Site location 	
Last maintenance  	
Last calibration 	
Checked by  	
Next maintenance due
Next calibration due 	
Figure 7.2.    Example of a gummed label
             for a high-volume sampler.
   4. Record all gasket replacements
 with dates and operator's initials in
 'the sampler log book.

 7.3   Rotameter
   1. Clean and recalibrate the
 rotameter of  a sampler when the float
 behaves erratically or when  moisture
 or foreign matter is detected in the
 rotameter.
   2. Clean the rotameter prior to
 routine calibration (alcohol is a
 satisfactory cleaning solvent).
   3. Refer to the flow diagram (Figure
 7.3) for the required maintenance
 steps.

 7.4   Sampling  Head
  Leaks in the sampling head occur
 infrequently. The welded seams and
the condition of the guide pins on the
top surface of the head should be
visually checked initially. Should  a

-------
                         Section 2.2.7
                                                                           Jan. 1983
 Figure 7.3.
              Disassemble
                Clean
                  \
                 ymii

                  I
  Examine
              Reassemble
  Calibrate


 Maintenance sequence for
rotameter.
defect be suspected, the following
procedure should be followed:
  1. Assemble the sampling head to
     the motor.
  2. Install a filter for resistance.
  3. Apply a soap solution to the
     suspect problem area.
  4. Disassemble the sampling head.
  5. Examine the inside of the head
     for soap bubbles.
  6. Repair or discard the sampling
     head if a leak is indicated by
     soap solution being inside of the
     head.

7.5  Motor Gaskets
  Two gaskets are used with each
sampler motor. The top rubber gasket
is approximately 3/16-in. thick and
the bottom foam rubber gasket is
approximately 3/4-in. thick.
  1. Inspect both gaskets for wear or
deterioration.
  2. Replace if necessary.

7.6   Flow Transducer and
Recorder
  Routine maintenance is not
required for this device. Should a
malfunction occur,  replace the old
recorder with a new one.
Table 7.7.    Activity Matrix for Maintenance"
Equipment
Sampler motor


Faceplate gasket

Rotameter

Motor gaskets

Sampling head
                  Acceptance limits
                     Frequency and method
                     of measurement
           Action if
           requirements
           are not met
                   400-50O h of motor brush
                   operation: no malfunction

                   No leaks at the filter seal

                   No foreign materials:
                   stable operation
                   Leak-free fit

                   No leaks
                     Visually check upon receipt
                     and after, each 400 h of
                     operation.
                     Visually check after each
                     sampling period.
                     Visually check at each
                     reading.
                     Visually check after each
                     400 h of operation.
                     Visually check after each '
                     4OO h of operation.	
           Replace 'motor brushes;
           perform other maintenance
           as indicated.
           Replace the gasket.

           Clean; replace if damaged.

           Replace gaskets.

           Replace sampling head.

-------
                              Jan. 1983
                                                                              Section 2.Z.8
                                        8.0   Auditing Procedure
    An audit is an independent
  assessment of the accuracy of data.
  Independence is achieved by having
  the audit made by an operator other
  than the one conducting the routine
  measurements  and by using audit
  standards and equipment different
  from those routinely used in
  monitoring. The audit should be a true
  assessment of the accuracy of the
  measurement process under normal
  operation—that is, without any special
  preparation or adjustment of the
  system. Routine quality assurance
  checks by the operator are necessary
  for obtaining good quality data, but
  they are  not part of the auditing
  procedure.
  • Three performance audits and one
  systems audit are detailed in
  Subsections 8.1 and 8.2. These audits
  are summarized in Table 8.2 at the
  end of this section. See Sections
  2.0.11 and 2.0.12 of this volume for
  detailed procedures for systems audits
  and performance audits, respectively.
   Proper implementation of an
  auditing program serves a  two-fold
  purpose:  to ensure the integrity of the
 data and  to assess the accuracy, of the
 data. A technique for estimating the  '
 accuracy of the  data is given in
 section 2.0.8 of  this volume.


8.1   Performance Audits
  Performance audits conducted by
another operator/analyst provide a
quantitative evaluation of the quality
of the data produced by the total
measurement system (sample
collection, sample analysis, and data
processing). Performance audits of
three individual portions of  the total
measurement system are
recommended:
  1. Flow rate calibration
  2. Exposed filter reweighing
  3. Data  processing.

8.1.1   Audit of Flow Rate Calibration -
The frequency of audits of the flow
rate depends on the use of the data
(e.g., for PSD3 air monitoring or for
SLAMS4).  It is recommended that the
flow rate of each high-volume
sampler be audited each  quarter. Any
type flow-rate transfer device
acceptable for use in calibration of
high-volume samplers may be used as
the audit flow-rate reference
standard; however, the audit standard
must be different from the standard
 used to calibrate the high-volume
 samplers. The audit standard must be
 calibrated with a positive-
 displacement standard volume meter
 (i.e.. Roots meter) traceable to the
 National Bureau of Standards. See
 Subsection 2.2 for procedures used to
 certify flow rate transfer standards.
   With the audit device in place, the
 high-volume sampler should be
 operated at its normal flow rate. The
 differences in flow rate (in std
 mVmin) between  the audit flow
 measurement (X) and the flow
 indicated by the sampler's normal
 flow indicator (Y) are used to calculate
 accuracy as described in Section 2.0.8
 of this volume.
   Great care must be taken in
 auditing high-volume samplers having
 flow regulators because the
 introduction of the audit device can
 cause abnormal flow patterns at the-
 point of flow sensing. For this reason,
 the orifice of the flow audit device
 must be used with a normal glass
 fiber filter in place (and without
 resistance plates)  in  auditing flow-
 regulated high-volume samplers, or
 other steps should be taken  to assure
 that flow patterns  are not disturbed at
 the point of flow sensing.
  Detailed procedures and forms used
 to perform flow rate  audits are given
 in Section 2.0.12 of this volume.

 8.1.2  A udit of Exposed Filter
Reweighing - To avoid possible loss of
volatile components,  exposed filters
should be weighed, including any
necessary reweighing, as soon after
collection and equilibration as
practical. Thus, it may be impossible to
have lot sizes of more than 10 or 20
exposed filters. The procedure is as
follows:
  1.  Select randomly and reweigh
     four re-equilibrated filters out of
     every group of 50 or less. (This
     would mean 100 percent
     checking if four  or fewer exposed
     filters were weighed at  one
     time). For groups of 50 to  100,
     reweigh 7 from  each group.
     These suggested starting
     frequencies may be altered,
     based on experience and data
     quality. Decrease the frequency if
     past experience  indicates that
     the data are of good quality, or
     increase it if the data are of poor
     quality.  It is more important to be
       sure that the sample is
       representative of the various
       conditions that may influence
       data quality than to adhere to a
       fixed frequency.
    2. Reweigh all filters in a lot if any
       audit weight differs by more than
       ±5.0 mg from the original
       weight.
    3. Accept the lot with no change if
       all.audits are within ±5.0 mg of
       the  originals.
    4. Record the original and the audit
       weights  in milligrams (mg) on an
       X-and-R chart (Figure 8.1). Plot
       the  difference (d), defined as:

       d = original weight - audit weight.
                          Equation 8-1
       Tentative warning and control
       limits of  ±3.3 and ±5.0 mg,
       respectively,  are recommended
       until sufficient data are obtained
       to support an alteration of these
       limits. Out-of-control points.
       indicate the need for
       recalibration  of the balance'
       and/or improved operator
       technique. Do not increase the
       limits; however, more stringent
       limits may be established if
       experience warrants.
    5.  Forward the X-and-R chart to the
       supervisor for review.
    6.  Reweigh all of the remaining
       exposed filters in the lot if the
       balance requires recalibration or
       the operation technique is
       changed.
 8.1.3 Audit of Data Processing. - For
 convenience, the data processing
 should be audited soon after the
 original calculations have been
 performed. This allows corrections to
 be made immediately. This also allows
•for possible retrieval of additional
 explanatory data from field personnel  '
 when necessary. The procedure is as
 follows:
  1. Use the audit rate of Subsection
     8.1.2.
  2. Starting with the raw data on the
     data form or on the flow rate
     recorder chart, independently
     compute the concentration (in fjg
     TSP/m3) and compare it with the
     corresponding concentration
     reported on the SAROAD form. If
     the mass concentration
     computed by the  audit check Oug
     TSP/m3)a does not agree (within
     round-off error) with the original

-------
                       Section 2.2.8
                                                                        Jan. 1983
   II
   
-------
                              Jan. 1983
                                              Section 2.2.8
     value (fjg TSP/m3)m. recheck all
     samples in the lot and correct
     them as necessary.
   3. Record the audit values in the
     data log, and report them along
     with the original vatues to the
     supervisor for review. The audit
     value is always given as the
     correct value,  based on the
     assumption  that a discrepancy
     between the two values is
     always double-checked by the
     auditor.

 8.2   Systems Audit
   A systems audit is an on-site
 inspection and review of the quality of
 the total measurement system (sample
 collection, sample analysis, data
 processing, etc.),  and it is normally a
       qualitative appraisal. The procedure is
       as follows:
         1. Conduct a systems audit on
            receipt of a new monitoring
            system and as appropriate
            thereafter to audit possible
            degradation or significant
            changes in system operation.
         2. Use the preliminary checklist
            given in Figure 8.2. Check the
            questions for applicability to  the
            particular local, State, or Federal
            agency.
         See Sections 2.0.11 and 2.0.12 of
       this volume for detailed procedures  -
       and forms for systems audits and
       performance audits, respectively.
 Table 8.2.     Activity Matrix for Auditing Procedure
Audit
Flow rate
Acceptance limits
Percentage difference.
Frequency and method
of measurement
Once each quarter
Action if
requirements
are not met
Recalibrate before
resuminq samolina.
Exposed filter reweighing
Data processing
Systems
     X
within 37%
Audit weight = original
weight 35 mg
Audit concentration agrees
with original reported con-
centration within round-off
error

Method described in this
section of the Handbook
Perform 7 audits/100
filters, or 4 audits/<50
filters; use analytical
balance; condition filters
for 24 h before weighing.
Independently repeat cal-
culation of JSP concentra-
tion from data record for 7
samples per 100 (minimum
of 4 per lot).
At beginning of a new
monitoring system and
periodically as appropriate, '
observe procedures and
use checklist.
'Reweigh all filters in the
lot.
                                                                                        Recheck all calculations.
                                                                                        Initiate improved methods
                                                                                        and/or training programs.

-------
                            Section 2.2.8                    4                      Jan. 1983
                                        Checklist for Use by Auditor for Hi-Vol Method



  1  What type of hi-vol samplers are used in the network?	
 2  How often ar-e the samplers run? la) daily fbj once every 6 days Id once every 12 days Id) other

 3  What type of filter and how many are being used?  	
 4  Are there any preexposure checks for pin holes or imperfections run on the filters?

 5  What is the collection efficiency for your fitters-'  	,	
 5  What is the calibration procedure for the hi-vol sampler'
 7  Which statement most closely estimates the frequency of flow rate calibration? (a) once when purchased (b) once when
    purchased, then after every sampler modification (cl when purchased, then at regular intervals thereafter  	
 3  Are flow rates measured before and after the samp/ing period?'


     Yes	  No 	1_


 9  Is there a loo. book for each sampler for recording flows and times? Yes	No
 10  Are titters conditioned before initial and final weighings?	If so. for how long?	At what
     percentage humidity? _	;_	;	


 / /  Is the balance checked periodically?	// so. how often?	With which standard weights? 	

 12  How often are the hi-vof filters weighed?  	:	        '	
    How are the data from these weighings handled?
 13 Are all weighings and serial numbers of filters kept in a log book at.the laboratory?
 14  What is the approximate time delay between sample collection and the final weighing? ___^_^_ days


Figure 8.2.    Example of simplified checklist for use by auditor for hi-vol method.

-------
                             Jan. 1383                    1                  Section 2.2.9
            9.0   Assessment of Monitoring Data for Precision  and Accuracy


 9.1   Precision
   For each monitoring network.
 collocate an additional sampler at a
 minimum of one site (two sites are
 required for SLAMS") as follows:
   1.  Select a site with the highest    .                  •            .
      expected geometric mean
      concentrations.
   2.  Locate the two high volume
      samplers within 4 m of each
      other, but at least 2 m apart  to        .-                                                   ..
      preclude air flow interference.
   3. 'Identify one of the two samplers
      at the time of installation as  the
      sampler for normal routine
      monitoring; identify the other as
      the duplicate sampler.
   4.  Be sure that the calibration,
      sampling, and  analysis procedure
      are the same for the collocated
      sampler as for all other samplers
      in the network.
   5.  Operate a collocated sampler
      whenever its associated routine
      sampler is operated.
   6.  Use  the differences  in the
      concentrations (+g TSP.'std m3)
      between  the routine and                                                      '
      duplicate samplers to calculate           .           -                                     .
     the  precision as described  in
   -  Section 2.0.8 of this Handbook.
   Based on the results of a
 collaborative test,5 percent difference
 (Equation 8-1 of Section 2.0.8) should
•not exceed 31 5%.* An example
 calculation is given  in Section 2.0.8 of
 this Handbook.

 9.2  Accuracy
   The accuracy of the high-volume
 method for measurement of TSP is
 assessed  by auditing certain portions
 of the measurement process, as
 described in Section 2.2^8. The
 calculation procedure for single
 instrument accuracy is given in    -   '
 Section 2.0.8 of this volume of the
 Handbook.
•This 315% is calculated at the 99 7 probability
interval This means that if the two samplers do
agree, chances are less than 3 out of 1000 that a
difference larger than I 5°'o will be observed

-------

-------
                              Jan. 1983                    1                Section 2.2.10
               10.0   Recommended Standards for Establishing Traceability
   For data of the desired quality to be
 achieved, two considerations are
 essential: (1) the measurement
 process must be in a state of
 statistical control at the time of the
 measurement, and (2) the
 combination of systematic  errors and
 random variation (measurement
 errors) must yield a su+tably small
 uncertainty. Evidence of good-quality
 data requires the performance~of
 quality control checks and
 independent audits of the
 measurement process;  documentation
 of the data on a  quality control chart;
 and the use of materials, instruments,
 and measurement procedures that
 can be traced to  an appropriate
 performance standard.
   Data must be routinely obtained  by
 repeating measurements of Standard
 Reference samples (primary,
 secondary, and/or working standards),
 and a condition of process  control
 must be established. The working
 calibration standards should.be
 traceable to standards c; higher
 accuracy, sjjc'h as those' listed here.

 10.1   Recommended
 Standards for Establishing
 Traceability
  1. Class-S weights of NBS
 specifications are recommended  for
 the analytical balance calibration. See
 Subsection 2,1 for details on balance
 calibration checks.
  2. A positive displacement
 rootsmeter is recommended for
 calibrating the flow rate transfer
 standard that is used to calibrate the
 high-volume sampler. See Subsection
 2.6 for details on high-volume
 sampler calibration.
  3. A positive displacement
 rootsmeter (including a resistance
 plate) is recommended for calibrating
the device used to audit the high-
volume-sampler flow-rate calibration.
 See Subsection 8.1  for details on
flow-rate  calibration audits.
  4. The elapsed-time meter,  checked
semiannually against an accurate
timepiece, must be within 32
min/day.

-------

-------

-------
"I	II      I    III

-------
                                              Section No. 2.3
                                              Revision No.  0
                                              Date July 1,  1979
                                              Page 1 of 5
                          Section  2.3

          REFERENCE METHOD FOR THE DETERMINATION OF
             NITROGEN DIOXIDE IN  THE ATMOSPHERE
                     (CHEMILUMINESCENCE)
                           OUTLINE
          Section
SUMMARY
METHOD HIGHLIGHTS.
METHOD DESCRIPTION
     1.
     2.
     3.
     4.

     5.
     6.
     7.

     8.

     9.
    10.
    11.
 PROCUREMENT OF APPARATUS
 AND  SUPPLIES
 CALIBRATION OF EQUIPMENT
 OPERATION  AND  PROCEDURE
 DATA REDUCTION, VALIDATION,
 AND  REPORTING
 MAINTENANCE
 AUDITING PROCEDURE
 ASSESSMENT  OF  MONITORING
 DATA FOR PRECISION AND
 ACCURACY
 RECOMMENDED  STANDARDS FOR
 ESTABLISHING TRACEABlLlTY
 REFERENCE METHOD
 REFERENCES
DATA FORMS
Documentation
2.3
2.3
2.3.1
2.3.2
2.3.3
2.3.4
2.3.5
2.3.6
2.3.7
Number of
Pages
1
3
8
27
10
5
2
12
1
2.3.8

2.3.9
2.3.10
2.3.11
 9
 1
17

-------
                                             Section No. 2.3
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 2 of 5
                           SUMMARY
     Concentrations of nitrogen  dioxide  (NO2)  in ambient air are
determined  indirectly by photometrically  measuring  the  light
intensity,  at  wavelengths greater  than  600  nm, resulting  from
the  chemiluminescent  reaction of  nitric oxide  (NO) with  ozone
(03).  NO -  is  first quantitatively  reduced.to NO by a converter.
The  NO,  which  commonly  exists  in  association  with  NO-,  passes
through  the  converter  unchanged,   resulting  in  a  total  NO
                                                                H
(nitrogen oxides)  concentration  of NO  + NO,,.    A portion  of the
ambient  air  is  also  reacted  with  O3  without  having  passed
through the converter,  and the  NO  concentration measured.   This
value  is   subtracted   from  the  NO   concentration  yielding  the
                                   X
concentration of NO_.
    -The.NO and the NO + NO2  measurements may.be made either con-
currently with a dual channel detection system or cyclically with
a single  channel  system  as  long  as the cycle time is not greater
than 1 min.
     References  1  and  2 were  used  extensively for the  method
description.  Reference  3 was used in the  development  of  effec-
tive quality assurance procedures.

-------
                                              Section No.  2.3
                                              Revision No.  0
                                              Date July 1,  1979
                                              Page 3  of 5
                       METHOD HIGHLIGHTS
      In this  quality assurance  document for  the NO-  Reference
 Method the  procedures are  designed to  serve  as guidelines  for
 the development  of agency  quality assurance programs.   Because
 recordkeeping is  a critical  part of quality  assurance  activities,
 several data  forms  are  included to  aid in--the  documentation  of
 necessary  data.   The blank  data forms  (Section 2.3.11) may  be
 used as they are,  or they may be used as guidelines  for preparing
 forms   more-  appropriate   to  the  individual   agency;   partially
 filled-in  forms   are  interspersed  throughout ,the  text  of the.
 method description  to illustrate their uses.  Activity matrices
 at  the end of pertinent  sections can be used for quick review  of
 the material covered in the  text sections.  Following  is  a brief
 summary of the material  covered in this N02 method description.'
 1-   Procurement  of Equipment   '                      -      •
  .   Section  2,3.1 gives  the specifications, criteria,  and design
 features of  the equipment  and material  required for the operation
 and quality  assurance of  a  continuous NO  analyzer.   The selec-
                                          X
 tion  of the correct  equipment and-supplies  is  a prerequisite  to
 a quality  assurance  program.   This section is designed  to  provide
 a guide for  the  procurement and initial  check of equipment and
 supplies.                                                    •
 2 -   Calibration of  Equipment
     Section  2.3.2 provides  procedures  and  forms to  be used  in
 the -selection  and  checking.of  calibration equipment, performing a
 multi-point  calibration,   and  evaluation of  calibration data.
 Subsections. 2.1, 2.2, and 2.3  deal primarily with minimum  accept-
 able requirements  for equipment  and  standards  to be used in the
 generation of NO2  concentrations.   Detailed procedures for the
 acceptance of NO2,  GPT  calibrators are  also  given.    Subsection
 2.4 provides  a step-by-step  description of the recommended cali-
bration  procedures  for an NO2  chemiluminescence analyzer along
with example  calculations.   The data  form  (Figure 2.1 of  Sec-

-------
                                             Section No. 2.3
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 4 of 5

tion 2.3.2)  is to be used in  documentation  of calibration data.
The  primary element  of  quality control  is  dynamic  instrument
calibration.
3.   Operation and Procedure
     Section 2.3.3  outlines  protocol  to 'be  followed  by  the
operator  during  each site  visit.   Checks should  include visual
inspection  of  the shelter,  sample  introduction system, analyzer
and recorder.  In addition analyzer performance checks consisting
of zero,  span,  and precision points are to  be made.   To provide
for  documentation  and accountability of  activities,  a checklist
similar to  the example provided  in Figure 3.1 of  Section 2.3.3
should 'be compiled and  then filled out by the field-operator as
each  activity  is  completed.    Analyzer Level  1  zero  and  span
checks must be carried  out  at least once every two weeks.  Level
2 zero and span checks should be conducted in between the Level 1
checks at a frequency desired by the user.   Span concentrations
for  &oth  levels  should  be between 70 .and  90%  of the measurement
range.  A one point precision check is to. be done every two weeks
at an  NO, concentration between  0.08  and 0.10  ppm.   Data forms
         £t
similar to Figures 3.2 and 3.3 of Section 2.3.3 are to be used in
documenting the  analyzer  performance checks.   An essential part
of the quality assurance  program is a scheduled series of checks
for the purpose  of verifying  the operational status, of the moni-
toring system.
4.   Data Reduction
     Section 2.3.4 describes procedures  to be used for editing of
strip  charts  and  for data reduction.   Data  collected on. strip
charts serve no  useful function until  converted into meaningful
units (ppm,  [ig/m ) by hourly averaging and application of a cali-
bration relationship.   These  data must then be transcribed into
an appropriate data' format  such as the SAROAD Hourly Data form.
5.   Maintenance
     Section 2.3.5  addresses  recordkeeping  and  scheduled activ-
ities pertinent to preventive and corrective maintenance.  A sam-

-------
                                             Section No. 2.3
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 5 of 5

pier maintenance log is presented in Figure 5.1 of Section 2.3.5.
Preventive  and  corrective maintenance  are  necessary to minimize
loss of  air quality data due to analyzer malfunctions and out of
control conditions.
6•   Assessment of Data for Accuracy and Precision
     Section 2:3.6  discusses  system and performance audits along
with audit procedures and forms.  Accuracy of data is assessed by
performing an independent audit.
     Multipoint performance audits used to assess the accuracy of
the data collection are discussed in Subsection 6.1.  Examples of
an audit summary form and audit calculation form are presented in
Figures  6.1  through 6.4.  Data  reduction audit is  discussed, in
Subsection 6.2 and a systems audit in Subsection 6.3.  Figure 6.5
presents an ' example checklist that  may be used by the auditor.
     Section  2.3.7 describes  the  techniques  for  assessment  of
accuracy and precision.                 "   .
7•   Reference Information
     Section 2.3.8  discusses  the  traceability  of measurements to
established  standards  of  higher accuracy,  a necessary  prereq-
uisite for obtaining accurate, data.
     Sections 2.3.9  and  2.3.10 contain the Reference  Method and
pertinent references.
     Section 2.3.11 contains blank data forms  for the convenience
of the user.

-------

-------
                                              section No. 2.3.1
                                              Revision No. 0
                                              Date July 1, 1979
                                              Page 1 of 8
                      METHOD DESCRIPTION
 1.0  PROCUREMENT OF APPARATUS AND SUPPLIES
      The measurement  of NO_ in  ambient air requires  basic sam-
 pling equipment  and other supplies.  These  include,  but are not
 limited to, the.following:
      1.    Reference  Method  NO2  chemiluminescent  analyz.er  (see
 Subsection 1.1 for an address for obtaining an up-to-date list of
 analysers).
      2.    Strip chart recorder or data logging system,
      3.    Sampling lines,
      4.  •  Sampling manifold,
      5.    Calibration equipment,
      6.    NBS calibration standard,
      7.    Working gas traceable to  NBS standard,
      8.'    Zero air (Reference 2),
     '9.    Spare parts,
     10.    Record  forms,  and  .
     11.    Independent audit system.
 Purchases of these  supplies  should  be recorded in a  log  book  to
 provide  a reference  for future procurement needs and  for future
 fiscal planning.   An example of this  log  is  Figure  1'. 1.   Quality
 assurance activities  for  procurement  of  apparatus  and  supplies
 are  summarized in Table  1.1 at  the  end of  this  section.
-1-1   Oxides- of Nitrogen  Chemiluminescent Analyzer
      Chemiluminescent  NOx  analyzers, currently available  for the
 measurement  of NO2  in  ambient  air,   are  competitively  priced.
 However,  price differences do  become  apparent when options are
 ordered.   Available options  range  from automatic zero and  span
 functions to complete telemetry systems that transmit daily  zero
 and  span checks  and real-time data  from  the  site  to a  central
 location.   Although these  options  have  advantages, their  absence
 from  the basic monitor  will  not detract  from  performance.  The

-------
    c
    OJ

    S
    o
    o
  I  C
  o o
  C.T-
  W ->
     u
  .
"
!- O =
=   c
    o-
      Si
      M
                                                Section No.  2.3.1
                                                Revision No.  0
                                                Date  July 1,  1979
                                                Page  2 of  8
 .
N
      n
      ^s ^
      3
                                                -p

                                                
-------
                                              Section No. 2.3.1
                                              Revision No. 0
                                              Date July 1, 1979
                                              Page 3 of 8

 necessity and  desirability of options will be  dictated by field
 personnel availability,  site accessibility,  and budget limita-
 tions.
      Only analyzers designated by EPA as  a reference or an equiv-
 alent method  should be  purchased.   When  purchasing,   the  buyer
 should  request that the manufacturer supply documented proof that
 the- analyzer  does  perform  within  specifications  (Table 4.1,
 Section 2.0.4).   The  best evidence  is  a  strip chart  recording
 showing the  specific analyzer's  zero  drift,   span drift,  elec-
 tronic  noise,  rise time,  fall  time, and lag  time.   -The  strip
 chart will  also  serve  as a reference to determine whether  the
 performance  of the  analyzer has  deteriorated at  a later date.   In
 addition,  the  user  should reverify these performance characteris-
 tics either  during  the  initial  calibration or by using abbrevi-
 ated forms of the  test  procedures provided in Ambient  Air  Moni-
 toring  Reference  and Equivalent  Methods.  40   CFR 53,  Federal'
 Register-,  Vol.  40, 'No.  33, pp.  7052-7060,  February  18,  1975.
      Acceptance  of  the  analyzer  should be 'based on results  from
 these performance tests.   Once accepted,  reference and  equivalent
 analyzers  are warranted by the manufacturer to operate  within  the
 required performance limit for one year.
      An  up-to-date  list  of analyzers designated.as reference' or
 equivalent methods 'for NO2  is available by writing to:
          U.S. Environmental Protection Agency
          Environmental Monitoring Systems Laboratory
          Department E, MD-77
          Research Triangle Park, North Carolina  27711
 1-2   Strip Chart Recorder
     Recorders  are  commercially available  in  a  wide  variety of
prices  and specifications.  Factors  to  be considered  when pur-
chasing  a recorder are:
     1.    Compatibility with  the  output  signal  of the analyzer,
     2.    Chart width  (a  minimum  of  15 cm  (6 in.) is recommended
for the  desired accuracy in data  reduction,
     3.    Chart speed (at least 2.5 cm (1  in.)  per hour),

-------
                                             Section No. 2.3.1
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 4 of 8

     4.   Response time,
     5.   Precision and reliability,
     6.   Flexibility of operating  variables  (speed,  range),  and
     7.   Maintenance requirements.
1.3  Sampling Lines and Manifold
     Sampling lines and manifolds should be constructed of Teflon
or  glass  to  minimize  possible- reaction  and degradation  .of  the
oxides of nitrogen.  The  residence  time within the sampling line
should be  minimized to reduce  the  possibility  of interreaction.
For example,  the  reaction  of ambient concentrations of NO and 03
in the sample lines and manifold would lead to erroneous measure-
ments .
1.4  Calibration Equipment
     To ensure accurate measurements of the NO and N02 concentra-
tions, calibrate  the  analyzer  at the time of  installation,- and
recalibrate it:
    "1.   no  later  than three  months after the most recent, cali- -
bration or performance audit which indicated analyzer response to
be acceptable; or
     2.   following any one of the activities.listed below:
          a.   an interruption  of more than  a  few days  in ana-
lyzer operation;
        •  b.   any  repairs  which might affect  its  calibration;
          c.   physical relocation of the analyzer; or
          d.   any other indication  (including  excessive zero or
span drift)  of possible significant inaccuracy  of the analyzer.
Following any of the activities listed in item 2 above,-a level 1
zero and span check should be made  to determine if a calibration
is necessary.  If the  analyzer zero  and span drifts do not exceed
the calibration limits  in  Table 9.1 of Section 2.0.9, Subsection
9.1.3,  a  calibration  need not be performed.  If either the zero
or span drift exceed their respective calibration limit, investi-
gate the cause of the  drift, take corrective action and calibrate
the analyzer.

-------
                                              Section No. 2.3.1
                                              Revision No. 0
                                              Date July I, 1979
                                              Page 5 of 8

      When purchasing  or  designing a calibration  system,  be sure
 that the calibration  system  meets  the  guidelines outlined in the
 revised Appendix F,  Federal Register,  Vol.  41,  No. 232,  p. 52688,
 December 1,  1976.   Calibration  procedures are  also  detailed in
 the Technical  Assistance  Document (TAD).2                       '
     .Two methods  for  dynamic multipoint calibration  of nitrogen
 dioxide analyzers are'  specified in the  Federal  Register.1
      1.   Alternative  A:   .Gas  phase   titration   (GPT)  of  an  NO
 standard with  O3 to  generate  known concentrations of NO_.
      2.   Alternative  B:   NC>2 permeation tube,  a  dynamic dilution
 •system   to  produce  known concentrations   of  NO-,   and -an  NO
 cylinder.
      Only alternative  A,  GPT,  is  discussed'  in this  document.'
 Those  wishing  to  use  alternative  B should read Section 2 of the
 TAD.    Alternative  A  requires four major  components.  They are.
      I-    Working NO gas  standard,
      2.    Stable O3  generator,    .                            '
      3.    Source of  zero  air,2  and
     4.    GPT  system.
 The  NO  standard  must be traceable   to  a  National 'Bureau  of
 Standards, Standard  Reference Material  (NBS-SRM).3  An acceptable
 protocol  to  demonstrate  the  traceability of commercial cylinder
 gas to  NBS-SRM cylinder gas is  described in Section 2.0.7  of this
 volume  of the Handbook.
     Zero  air  (free  of contaminants that  can cause a detectable
 response  with  the oxides  of  nitrogen analyzer  or that can  react
 with either  NO,  q>3,  or NO2) is commercially available, or it can
 be  generated by  the user.   Detailed  procedures  for generating
 zero air are in TAD.2
     .The equipment that is needed to carry out the calibration  is
 commercially available, or it can be assembled by the user.   When
purchasing a calibrator  or its components,  certain factors  must
be considered.
     1.    Traceability of  the certified calibration  gases to  an
NBS-SRM.

-------
                                             Section No. 2.3.1
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 6 of 8

     2.   Accuracy  of  the  flow-measuring device (rotameter, mass
flow meter, bubble meter).
     3.   Maximum and  minimum flows of dilution air and calibra-
tion gases.
     4.   Stability of the 03 generation.
     5.   Ease of transporting  the calibrator from site to site.
As  a precaution,  all  new  GPT  apparatus  should be  checked  out
against a calibrator of known reliability..
1.5  Spare Parts and Expendable Supplies
     In addition to the basic  equipment discussed above,  it is
necessary to maintain  an inventory of spare parts and expendable
supplies.    The  manufacturer's  manual  contains a  section  de-
scribing  the parts  that  require  periodic  replacement  and  the
frequency  of  replacement.   Based  on  these  requirements,  the
management  of  the  monitoring network can determine  which parts
and  the quantity of each that should  be available  at all times.
A  generalized  list  of  spare parts  and expendable  supplies  is
provided below  (for specific requirements refer to the manufac-
turer's manual):
     1.   Particulate filters,
     2.   Sampling lines,
     3.   Pump diaphragms,
     4.   Dryer columns,
     5.   Activated charcoal,
     6.   Recorder chart paper,
     7.   Recorder 'ink or pens,
     8.   Calibration gas,
     9.   Record forms, and
    10.   Spare fittings  and glassware assortment.
1-6  Reanalysis of Calibration Working Standards
     All working  standards used  for  calibration purposes should
be reanalyzed by the, user at least once every 6 mo.
     Section 2.0.7  (in particular,  Subsections  7.1.2  and 7.1.5)
describes  the procedures  for analysis and for reanalysis of

-------
                                              Section No.  2.3.1
                                              Revision No.  0
                                              Date  July 1,  1979
                                              Page  7  of 8

 cylinder  gases).   Flow-measuring  devices  should be  recalibrated
 by  following  the procedures  and  schedules  in  Section 2.1.2.
 1.7 Record Forms
     Record  keeping is  a  critical part of all quality assurance
 programs.   Standard forms similar to those  that  appear  in this
 manual should be developed for individual  programs.   Three things
 to  consider in.the development of  record forms are:
     1.   Does the form  serve a  necessary  function?
     2.   Is  the documentation complete?
     3.   Will the  forms be' filed in such a manner  that they can
 easily be retrieved when needed?            :
 1-8  Audit Equipment
     Personnel,  equipment,  and  reference  materials  used  in con-
 ducting audits  must be  independent  from  those normally  used in
 calibrations  and operations.
     Known concentrations  of N02 can be generated by the GPT of
NO with 03 to produce NO2 or by the use-'of an NO2 permeation tube
 and a  dynamic dilution  system.  All  audit gas  standards  must be
traceable to  NBS-SRM's  as  described by  the  protocol  in  Sec-
tions 2.3.2  and  2.0.7.  All  flow  rates  should  be  measured using
a calibrated  soap .bubble meter  or an equivalently accurate  pro-
cedure.

-------
                                                        Section No.  2.3.1
                                                        Revision  No. 0
                                                        Date  July 1, 1979
                                                        Page  8 of 8
              Table  1.1
          ACTIVITY  MATRIX FOR  PROCUREMENT
          OF  EQUIPMENT AND SUPPLIES
Equipment/
 Supplies
Chemiluminescent
 analyzer
Recorder
Sample lines
 and manifold
Calibration
 equipment
Working stan-
 dard NO cylin-
 der gas or N0_
 permeation
 tube
Record forms
Audit equip-
 ment
Acceptance limits
Meets performance
specifications in
Table 4.1, Sec 2.0.4
Compatible with output
signal of analyzer;
chart width of 150 mm
(6 in.) is recommended
Constructed of Teflon
or glass
Meets guidelines of
reference 1 and
Sec 2.3.2
Traceable to NBS-SRM;
meets limits in trace-
ability protocol for
for accuracy and sta-
bility (Sec 2.0.7)
Develop standard forms
Must not be the same
as used for calibra-
tion
Frequency and method
   of measurement
Manufacturer should
provide a strip chart
recording the specif-
ic analyzer's per-
formance; reverify
performance speci-
fications at initial
calibration
Check upon receipt
Check upon receipt
See Sec 2.3-9
Analyzed against an
NBS-SRM; see proto-
col in Sec 2.0.7
N/A
System must be
checked out against
known standards
Action if
requirements
are not met
Manufacturer
should make
proper adjust-
ments and re-
run the per-
formance check
Return equipment
to supplier
Return equip-
ment to supplier
Return equipment/
supplies to sup-'
plier
Obtain new work-
ing standard and
check for trace-
ability
                                                                Revise forms
                                                                as appropriate
Locate problem
and correct
or'return to
supplier

-------
                                              section No.  2.3.2
                                              Revision No.  0
                                              Date  July 1,  1979
                                              Page  1  of 27
 2.0   CALIBRATION OF EQUIPMENT
      The  accuracy  and  precision  of data  derived  from air moni-
 toring instrumentation are dependent on the  quality  control pro-
 cedures used,  primarily the dynamic  instrument  calibration.  Cal-
 ibration  determines the relationship between  the observed  and the
 true  values  of the  variable being measured.
      Dynamic  calibration  involves  introducing  gas   samples  of
 known concentrations into an  instrument in  order  to adjust the
 instrument  to  a predetermined sensitivity and  to produce  a cali-
 bration   relationship..   This  relationship is  derived  from  the
 instrumental response  to  successive samples  of  different known
 concentrations.   These  standard gas  mixtures may be introduced in
 a  decreasing order  of concentrations to minimize response times.
 As a  minimum,  three reference points and  a zero point are recom-
 mended...^ define  this.relationship.  The true values  of the cali-
 bration gas  must  be  traceable to NBS-SRM's (Section 2.0.7). -
      Most present-day monitoring instrument  systems  are subject
 to  drift and  variation  in internal  parameters  and cannot  be
 expected  to maintain  accurate  calibration over long periods  of
 time.  Therefore, it is necessary to dynamically check the cali-
 bration relationship on a predetermined  schedule.   Precision is
 determined  by  a  one-point check  at least once every two weeks.
 Accuracy  is  determined  by  a  three-point audit  once each quarter:
 Zero  and  span checks  must  be used to document  within-cpntrol
 conditions,   these  checks  are  also  used  in  data  reduction  and
 validation.   Table  2.1  at  the  end of this section summarizes the
 quality assurance activities for calibration.
 2.1  Calibration Gases
 2-1-1  Compressed NO in Nitrogen  -  The NBS-SRM's provide  refer-
 ences  against  which   all  calibration   gas   mixtures  must  be
compared  (Section 2.0.7).   The  steps  required to  compare  the
concentration  of  a   commercial,  working calibration  standard  to
an NBS-SRM   are  described in  Subsection  7.1  of  Section 2.0.7.

-------
                                              Section No.  2.3.2
                                              Revision No.  0
                                              Date  July 1,  1979
                                              Page  2  of 27

 Subsections  7.1.4  and  7.1.5 .describe  the  verification  and re-
 analysis  of  cylinder gases.
 2.1.2   NO^ Permeation Tubes  -  The steps required to  compare the
 concentration  of a commercial working calibration standard  to  an
 NBS-SRM are described in  Subsection 7.3.3 of Section 2.0.7; see
 Subsection 7.3.6 for the reanalysis  pf permeation  tubes.
 2.2  Dilution  Gases                   •     "
              2
     Zero air  (free of contaminants that could cause a  detect-
 able  response with  the  oxides  of nitrogen  analyzer  or that can
 react  with either NO, O3/  or NO2) is commercially  available,  or
 can  be generated  by the  user.   Detailed procedures  for gener-
 ating  zero air are in TAD.2
 2.3  Dynamic Multipoint Calibration  Principles
     Two  methods  for dynamic multipoint calibration  of nitrogen
 dioxide are specified in the  Federal Register.1
    JL.    Alternative A:    Gas  phase  titration  (GPT) of  an  NO
 standard  with  O3 to  generate  known concentrations  of NO_.
     2.    Alternative B:   NO-  permeation  tube  and  a  dynamic
 dilution  system to produce known concentrations  of NO? .
 Both  methods  provide reliable  results  when correct  calibration
 procedures are followed.  Experience has shown,  however, that NO._
 permeation tubes  may become  unreliable  if not handled properly."
 Furthermore, the conditions tha-t contribute to the degradation  of
 the tubes  are  not well understood  at this  time,  so care should  be
 exercised  by  those  using  alternative   B  for  calibrating NO
 analyzers.  Analyzers that require calibration of NO or NO   chan-
 nels must use an  NO standard  and a dynamic  dilution system  to
 generate known concentrations.  Both alternatives require the use
 of  an  NO calibration  gas  to  determine  the efficiency  of the
 analyzer's  NO2 to  NO converter.   Only  alternative  A  (GPT)   is
 discussed  in  this  document;  those using  alternative  B  (NO
 permeation tube)  should refer to TAD.2
 2.3.1   Gas Phase Titration  (GPT)  - The   principle   of  GPT   is
based  on  the  rapid  gas  phase  reaction  between  NO  and 0_ which

-------
o

                                              Section No. 2.3.2
                                              Revision No. 0
                                              Date July 1, 1979
                                              Page 3 of 27

 produces stoichiometric quantities of NO, as shown by the follow-
                                          N02 + 02.
 Given the NO  concentration is  known for this  reaction,  the re-
 sultant concentration  of NO2  can be determined.   Ozone  is  added
 to excess NO in  a  dynamic calibration system,  and the NO channel
 of  the  chemiluminescent  analyzer   detects  the  changes  in  NO
 concentration.   A'fter  the addition  of O3, the  observed  decrease
 in NO concentration on the calibrated NO channel is equivalent to
 •the concentration  of  NO2 produced.   The amount of  NO  generated
 may be varied- by adding  varying  amounts of O3 from  a stable
 generator.
      Dynamic  calibration systems  based  on  this  principle  are
 commerically  available,  or may be  assembled  by  the  user.   A
 recommended  calibration  system  is   described   in  the   Federal
 Register   and  detailed in TAD.2   Persons desiring to  assemble
 their own calibration system's  should  follow the procedures  in
 TAD.2
      Both  the  assembled  and  the purchased  calibration  systems
 must  meet  the  following conditions  before  being  used  for  NO'
 calibrations.
      1.   Use an NO standard gas traceable to an NBS-SRM.
      2.   Have  a stable  O3  source  with an  adjustable  output.
      3.   Have  a minimum  total flow output that exceeds the
 analyzer flow demand by at least 10%.
      4.   Be capable  of  generating  an NO concentration  that  is
 approximately  9'0%  of  the  upper  range   limit  (URL)  of  the  NO
 range to be calibrated.
      5.   Have a reaction chamber residence time of  <2 min.
      6.   Have a dynamic parameter specification of ^2.75 ppm-min
 at  the operating  conditions  at which  the  calibration  will be
performed.
      It has been  determined empirically  that  the NO-O  reaction
goes  to  completion  (<1% residual O3)  if  the NO  concentration in

-------
                                             Section No. 2.3.2
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 4 of 27

the reaction chamber  (ppm) multiplied by the residence time (min)
of  the reactants  in the  chamber is  >_2.75 ppm-min.   The theory
behind  the  development  of  this  equation  is  in  the  Federal
        1             2
Register  and in TAD.
2.3.2   GPT  Calibrator Check Procedure  - The following procedures
and equations should  be used to determine whether an existing GPT
calibration  system will meet  required conditions for a specific
calibration.
     For  calibrators that  have  known pre-set flow  rates,  use
Equations 2-5  and 2-6  of  steps 7 and 8  to -verify  the required
conditions.   If  the calibrator  does  not meet  specifications,
follow' the  complete  procedure to determine what  flow modifica-
tions must be made.
     1.   Select an NO standard gas that has a nominal concentra-
tion in the  range of 50 to 100 ppm.   Determine the exact concen-
tration  [NO]STDa  by referencing against  an  NBS-SRM,   as  dis-
cussed  in Section 2.0.7.
     2.   Determine  the  volume (cm )  of  the  calibrator reaction
chamber (VRC)-   If the actual volume  is  not  known,  estimate the
volume by measuring the approximate dimensions of the chamber and
using  an  appropriate formula  such  as  V  =  ^nr  for a  sphere or
      2                       "
V = rtr  1 for a cylinder.  The reaction chamber should riot be con-
fused with the mixing chamber where the dilution air and the gen-
erated NO2 are mixed.
     3.   Determine  the  required minimum total  flow output  (FT)
using Equation 2-1,
     F  = analyzer flow demand (cm /min) x  i'.      Equation 2-1
If more than  one  analyzer is to be calibrated  at the same time,
multiply FT by the number of analyzers.
<§
 Throughout this method description the notation [NO] will denote
 the concentration of NO; similarly for [NO_] and [NO ].

-------
                                             Section No. 2.3.2
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 5 of 2.7

     4.   Calculate  the  NO  concentrations    [NO]     needed  to
approximate 90% of  the  URL of the NO2 analyzer to be calibrated,
using Equation 2-2,

           '•N0^OUT = URL of analYzer (ppm) * 22_.    Equation 2-2
                OUT                             _.

      5.    Calculate the  NO flow  (FNO)  required to generate  the
 NO  concentration [NO]OUT/  using Equation 2-3,

                        _  [NO]OUT X FT
                    FNO  -- '•  TNOT - '         '      Equation  2-3
                             L   JSTD
      6.    Calculate the  required  flow  through the  ozone gen-
 erator  (FQ),  using Equation 2-4,'
                F  -X  FNO X VRC    ^
                *0 -]/     2.75.ppm-min     ~  FNO'       Equation  2-4
     7.   Verify  that the  residence time  (tR)  in  the  reaction
chamber is <2 rain, using  Equation 2- 5,
                       _    RC
                    tR ~ F—+~F— -2 min-            Equation 2-5
                          O   rNO
     8.   Verify that the dynamic parameter specification  (P ) of
the.  calibrator's   reaction   chamber   is   I2.75ppm-min  using
Equation 2-6,
Note:  if tR is >2 min or if PR is <2.75 ppm-min, changes'in flow
conditions (FT/ FQ, FNQ) or in the reaction chamber volume (V  ),
or both will have  to  be made, and tR  and  PR will have to be^e-
calculated.                              '
     9.   After Equations  2-5 and  2-6 are  satisfied,  calculate
the diluent air flow (FD) using Equation 2-7,

-------
                                              Section No .  2.3.2
                                              Revision No.  0
                                              Date  July 1,  1979
                                              Page  6  of 27

                 FD  =  FT  "  F0  ~  FNO'                   Equation 2-7

 2.3.3   Example Calculation - Following is  an example calculation
 that can be used to  determine  whether an existing GPT calibrator
 will meet  the required conditions  for a  specific  calibration.
 For  this  example,  it  is  assumed  that only the volume  of the
 reaction chamber,  VRC/  and the concentration of the NO standard,
 [NO]STD,  are known.   All flow settings (FNQ,  FQ/ FT/  and FD)  will
 be  calculated.  In many uses,  these  flow  settings  are known and
 need only to  be substituted in Equations  2-5  and  2-6 to verify
 the  required  conditions.   Before doing any calculations,  the URL
 and  flow demand of the  analyzer being  calibrated must be known.
 Operating parameters  are  determined  from  the operations  manual:
          Upper  range limit =0.5 ppm, and
                 Flow  demand = 2500 cm /min..
 Volume  of calibrator reaction  chamber  is  determined by physical
 measurement;

          VRC  =  180 cm3.

     The  concentration  of  the. NO standard  gas  to  be used is de-
 termined by reference against an NBS-SRM (Section  2.0.7),

           [NO]STD =50.5 ppm.

     1.   Determine the  minimum total flow (F_)  required at the
output manifold using Equation 2-1,
         FT = 2500 cm3/min (iI2) = 2750 cm3/min. .
Because  low  flows  are  difficult to  control and measure,  it is
often advantageous to set a higher total flow than needed.
     2.   Determine  the  highest NO  concentration,   [NO]    ,  re-
quired at the output manifold, using Equation 2-2,

-------
                                             Section No.  2.3.2
                                             Revision No.  0
                                             Date July  1,  1979
                                             Page 7 of  27
                                 Qf)

              [NO]OUT = °*5 ppm (} = °-45
     3.   Calculate the NO flow (FNQ) required to generate the NO

concentration [NO]QUT, using Equation 2-3,


            v   - 0-45 ppm x 2750 cm3/min _ OA c ,,3 ,  .
            FNQ --  50. -5 ppm - -- 24-5 cm /min-


     4.   Calculate . the require^ flow rate of the ozone generator

(FQ) using Equation 2-4,
     L
r   -1/5
•O  -y
                         •     3
            0.5 ppm x  24.5  cm /min  x  180  cm    Oy1  c    3 ,  .
                    2.7      -  - --  24'5  cm /min
                    2.75 ppm-min
         =  /80984 cm6/mih2 - 24.5 cm3
                                                       3
             '80984 cm /min   -  24.5  cm  /min =  260.08  cm /min.


     5.   Verify  that the  residence  time (tR)  in  the  reaction

chamber is <2 min. using Equation  2-5,


                             3
      >  _ ^___	  180 cm               «  /-^   •
      t-.. - 	=	:	=	  = Q.63  mi-n.
           260.08 cmj/min'•+  24.5  cm /min


     6.   Verify  the dynamic parameter specification  (P_)  of the
    •                                                    ^
calibrator  reaction  chamber,  using Equation 2-6  and  previously

determined, values,




    'PR = 50.5 ppm x 	'.	24 5  cm3/min	

                     260.. 08  cm /min +  24.5 cm3/min




           v 	 180 cm                 ^ „,_
                      3	5—:	 =2.75 ppm-min.
             260.08  cnT/min  + 24.5  cm  /min


     7.   Calculate  the   diluent  air  flow (FD)  required at  the

mixing chamber,  using Equation 7,



   FD = 2750 cm3/min - 260.08 cm3/min  - 24.5  cm3/min =  2465 cm3/min

-------
                                              Section  No.  2.3.2
                                              Revision No.  0
                                              Date  July  1,  1979
                                              Page  8 of  27

2.4   Calibration Procedures
      The  procedures  for  multipoint calibration of  an  oxides of
nitrogen  analyzer  by GPT of an NO  standard with 0_ are specified
                        1
in the Federal Register.   To  facilitate these procedures, opera-
tional  and calculation  data  forms, have been  developed.  These
forms  will aid in  conducting  a  calibration  and in providing for
the   quality  assurance  checks.    Detailed  descriptions   of  the
calibration ' theory  and  procedure's  for GPT  are  in  the   Federal
        1             2                                     	
Register   and in TAD.
     Documentations  of 'all data on  the  station, instrument, cali-
brator, - reference  standard,   and  calibration procedures  are of
prime  importance  since  the validity of the 'data collected by the
instrument is dependent on its calibration.
2.4.1.  General Calibration Recommendations - Calibration  must be•
performed  with  a calibrator that  meets all  conditions  specified
in Subsection 2.3.2.  Flow settings  (with the exception of the NO
flow," FNQ) and  the  concentration  of  the NO standard,  [NO]STD,
used  in  the GPT' calibration  for NO2 must be the same as those
used in the calculations of specified conditions.
     The user should be  sure that all  flow meters are  calibrated
under the  conditions  of use against a reliable standard such as a
soap  bubble  meter or wet test meter.   All volumetric  flow rates
should be  corrected  to  25°C  (78°F)  and  760 mm  (29.92 in.). Hg.
Calibrations of flow meters are discussed in TAD2.
     Precaution must be taken to remove 02 and other contaminants
from  the  NO pressure regulator and  the delivery  system prior to
the start  of calibration  to  avoid .any conversion of the standard
NO to NO2.  Failure to  do  so can cause significant  errors in
calibration.  This problem may be minimized by:
     1.   Carefully  evacuating  the  regulator,  when  possible,
after  it  has  been  connected to  the cylinder  and before   opening
the cylinder valve;
     2.   Thoroughly  flushing  the   regulator and  the  delivery
system with NO after opening the  cylinder Valve;

-------
                                             Section No.  2.3.2
                                             Revision No.  0
                                             Date July  1,  1979
                                             Page 9 of  27

     3.-  Not  removing the  regulator from  the cylinder  between
calibrations unless absolutely necessary.
                                                      o
Further discussion of these procedures is given in TAD  .
2.4.2   Calibration Procedure for NO and  NO ,  - The  GPT  requires
                                           J\
the use of the NO channel of the analyzer to determine  the amount
of  NO2 generated  by titration.   Therefore,  it  is  necessary "to
calibrate  and  determine the  linearity  of the  NO channel before
proceeding  with the  NO2  calibration.   In some  analyzers it  is
also necessary to calibrate  the NO  channel.  This  can  be done
                                    ^t
simultaneously with  the NO  calibration.   The following procedure
uses the calibration data form,  Figure 2.1, to aid in the  collec-
tion and documentation  of  calibration data.  During the  calibra-
tion,  the  analyzer  should  be  operating in its  normal  sampling
mode,  and  the  test  atmosphere  should pass  through  all  filters,
scrubbers,  conditioners., and  other components used during normal
ambient sampling  and as much of the  ambient  air  inlet system  as
is  practicable-.   All  operational  adjustments  to. the.  analyz.er
should be completed prior to the calibration.
     1.   Record the  station  name  and address on the calibration
data form..   Identify  individual  stations by their official names
and  addresses.   Where appropriate,  station   name,  and   address
should be  the  same  as those appearing on the SAROAD site  identi-
fication form  for  that  station.   This will help to eliminate any
confusion by persons not familiar with the station.
     2.   Identify the  person performing the calibration  and re-
cord the-date of calibration.
     3.   Identify the  analyzer  being calibrated.  The manufac-
turer's name, model,  and serial number should be recorded.
     4.   Identify the  calibration  apparatus  used.   If the cali-
brator was purchased, record  the manufacturer's name,  model, and
serial  number.   Calibrators, assembled  by  the  user   should  be
assigned an  identification number  so' that  calibrations  can  be
referenced to that particular apparatus.
     5.   Identify,  by  supplier  and cylinder number,  the refer-
ence standard  to be  used.   Record the concentration of the cali-

-------
                                                         Section No. -2.3.2
                                                         Revision  No.  0
                                                         Date  July 1,  1979
                                                         Page  10 of 27
2.

3.
6.


7.
      Station
Calibrated  by

Analyzer mfgr.
                    S7C/?7£
                     SJo/AJo,///a
     NO  reference standard
       Supplier
       Concentration [N0]0  ''
                         Std
       NO-  impurity,
 Reference to NBS-SRM
 By 	
Zero knob  setting
Span knob  setting

Temperature
                                               Date
                          - 79
4.  Calibrator  mfgr.
                Model
                S/N
                                                               G-P-r
                                  Cylinder number	^	
                                  Cylinder pressure   /s-p o jt?S~/'<3
                                  o                         ^^
                                /L<9-
                                           Date
                                     NO
                                          NO     NO
                                     Barometric pressure
                  Equations used for NO/NO3;/NO  calibration
    Equation 2-8


    WOUT - F7C
             fNO

    Equation 2-9
    (NO
    '
    Equations 2-10,  -1QA, -10B

    Response {% scale) =  [NO1OUT  100 + Zv,^.
                          URL          N0

    For N02 substitute [NO2]QUT and ZNQ .


    For NO  substitute [NO l_,_ and Zwn .
                                  "NO.
    Equation 2-11
                      - [NO]
                            REM
                                 FNO + F0 + FD
    If  tNO2'lMp ~ ° use Equation 2-11A.

    Equation 2-11A
                                            FNO =• f-low rate of NO  standard.

                                            FQ = flow rate of air  through 0,
                                                 generator.               •

                                            FD = flow rate of dilution air.
             .concentration at the
              output manifold.

      [NO]-T  = concentration of the
               undiluted NO standard.

      [NO2lIMp = concentration of NO_
                impurity in the
              .  standard NO cylinder.

      URL = upper range limit.

     • ^NO-'oRIG = concentration of NO
                before 03 is added
                during GPT.

        'REM

               during GPT.
                                                     after O3  is added
                                            ZNO' ZNO '  2N00  = recorder re-
                                                   x     2   sponse to
                                                            zero air.
      Figure 2.1   Example  of  a calibration data  form  (front  side)

-------
                                              Section NO. 2.3.2
                                              Revision No.  0
                                              Date  July 1,  1979
                                              Page  11  of, 27
           NO/NO  CALIBRATION AND  LINEARITY  CHECK
                j\,
Calibration
points NO/NO
X
Zero
80% URL-
1
2


1
F + F
D 0'
3, .
cm /mm
£720'

3-3-


(NO]OUT,
ppm
0.0
0.'/-fiS'
0. 3.0O
o. /oo


4
NO
recorder,
% scale
S. 0
%t. o
¥$. o
If. 0


[N°xW
ppm
e> o
0. &S"
0. 3. oo
O . Joo


. 6
NO
recorder,
% scale
s:o
ft. o
vs~. o
zz'. o


                   NO2 CALIBRATION BY GPT
Calibration
points NO
Zero
ORIG
80% URL
1
2


7'
[N0x],
ppm

^-^5"
0. . /O


12 •
N02
recorder'
% scale
^. o

ff. o
4S<5~. O
15. 0


Figure 2.1.  Example of a calibration data form  (backside).

-------
                                              Section No. 2.3.2
                                              Revision No. 0
                                              Date July 1, 1979
                                              Page 12 of 27

 bration gas and the amount of NO2 impurities for each cylinder as
 determined by the user.  Do not use cylinders with pressures <200
 psig for  calibration.   It has been shown that,  for some concen-
 tratioiis,   gases  in cylinders  become unstable  at  low  pressures
 (Section 2.0.7).   Provide  a record  of NBS-SRM  traceability  for
 any cylinder used in  a calibration,  and include' the date and the
 name of the person who conducted the referencing.
      6.   Record the  zero  and the  span knob settings  after  the
 calibration is complete  so  that these settings can be  used at a
 later  date  to determine  changes in  the instrument  performance
-characteristics..  Note:   Some analyzers  may have  separate zero
 controls for NO,  NOx/  and NO2/-  others may have separate zero con-
 trols only  for  NO and NO  ;  still others may have  only one zero
                          X
 control common to all  three channels.
      7.   Record the shelter temperature  and  barometric pressure
 at the  time of calibration.
    ,  8.   Use the, NO/NO v  part of. the data  form  (Figure 2.1)  for
                        "
 the systematic recording of data determined  during 'calibration of
 the NO  and NOX channels  of  the analyzer.  Because zero and cali-
 bration adjustments  differ between analyzers,  the manufacturer's
.manual  should be  consulted  before calibration is  performed.
           a.    Auto  ranging  analyzers  should be  calibrated  on
 all ranges  which are  likely to "be  used.   Select  the  operating
 range of the  analyzer to be calibrated.  Precision and accuracy
 for NO2 calibration are best obtained when  all  three  channels of
 the analyzer are  set to the  same  range.
           b.    Connect the recorder  output  cable(s) of  the ana-.
 lyzer to  the  input  terminals  of the  strip  chart recorder (s') .
 Make all  adjustments  to the  analyzer  based  on  the  appropriate
 strip chart readings.   Analyzer responses  in the  procedures  given
 herein  refer  to recorder  responses.
           c.   Adjust  flows  of  the diluent air and the  0_  genera-
 tor air to obtain the flows determined in  Subsection 2.3.2.   Be
 sure that  the  total air flow   exceeds  the  total  demand of  the

-------
                                              Section No. 2.3.2
                                              Revision No. 0
                                              Date July 1, 1979
                                              Page 13 of 27

 analyzer (s)  connected to  the output manifold so  that no ambient
 air will be  pulled into  the manifold vent.   Record the sum of the
 flows  of diluent air  (FD)  and O3  generator air  (FQ)  in column 1
 of Figure 2.1.
           d.    Allow the analyzer  to sample zero air until stable
 NO,  NOX/  and N02  responses are  obtained.   After  the  responses
 have, stabilized,  adjust  the analyzer zero  control(s).   (Offset-
 ting the analyzer zero adjustments to +5% of scale is recommended
 to facilitate  observing the  negative zero  drift. )  Record  the
 stable  zero  air responses under column  4 for NO and column 6 for
 NOx.   Record the  NO2  zero  air  response  in  column 12,  Figure 2.1.
           e.    Adjust  the NO  flow from the standard  NO cylinder
 to generate  an NO  concentration of approximately 80% of  the URL
 of the  NO channel.   Measure  the NO flow  (FNQ)  and record it under
 column  2  on  the SO'% URL  line.
           f.    Calculate  the   exact  NO  concentration   [NO]Q  ,
 using Equation  2-8,    '
          fNOl      FNO X
          L   JOUT ~ F - +~~F - + F  '                    Equation  2-8

          Calculate the exact NO  concentration  [NO  ]nTTrr, using
      .                          "          -        X UU -L
Equation 2-9,
   [NO 1    -     X (IN°3STD +                       f   „.
   L  xJOUT ~      F   ITT — +~F - : - •            Equation 2-9

Record the [NO]QUT .under column 3 and the  [NOx]QUT under column 5
on the 80% URL line.
          g.    Sample  the  generated  concentration until  the NO
and the  NOx  responses have stabilized.  Adjust  the  NO span con-
trol  to  obtain  a  recorder response  as determined  by Equation
2-10,
Recorder response (% scale) = f  URgUT x 100 U ZNQ .  Equation 2-10

-------
                                             Section No. 2.3.2
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 14 of 27

where  URL = nominal upper  range limit  of the  NO  channel being
calibrated,  ppm.   Note:   Some analyzers may  have  separate span
controls  for NO, NO  , and NO,; others may have separate span con-
                   *»        £+
trols  only for NO and NO ; while still  others may  have only one
                         ,X
span control common to all three channels.  If only one span con-
trol  is  available, make  the  span adjustment  on the NO. channel.
When  adjusting the  analyzer's NO  span  control,  substitute the
                                  J^
[NOX]QUT  and the  ZNQ 'in Equation 2-10 to determine the recorder
response.   If  substantial  adjustments  of the  span  controls are
necessary, recheck the zero span adjustments by  repeating steps f
and g.   Record the  NO" recorder response  under  column 4 and the
NO  recorder response under column 6 on the 80%  URL line.
  «Jt
           h.   After the  zero and the  80% URL  points .have been
set,  determine two  approximately evenly spaced points  between
zero  and  80% URL  without  further adjustment  to the instrument.
These-additional points can be generated by either increasing the
dilution  flow  (FD) or by decreasing the FNQ.  For each concentra-
tion  generated,  calculate  the .exact NO and  NO  concentrations
                                                 X
using Equations 2-8 and 2-9.  Record the required information for
each point under  the appropriate  column in the NO/NO  table in
Figure 2.1.
     9.    Plot  the- analyzer  responses,  expressed in %  chart at
the NO recorder  
-------
                                                 Section No.  2.3.2
                                                 Revision No.  0
                                                 Date July 1,  1979
                                                 Page. 15 of 27
     100
CJ
z
'Q
UJ
O
                                             Fi tted calibration
                                                 1 ine
                                   Tolerance  limits for  instrument
                                      1inearity check, +2%
  Figure 2.2.   Example of an  NOx calibration  relationship.

-------
                                             Section No. 2.3.2
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 16 of 27
Calibration
point
Zero
80% URL
1
2

•
Concentration ,
ppm
X
o
a. 40
#. 2o
O. /o


x2
O
0./L
d.O*!
0.0}


Recorder
reading,
% scale
y
JT
.0
-?.r


Zx =
Ix2 =
,  Zy =
                                ,  Iy2  =
                            ,  Zxy =
                                                             5"
„ x = Ix/n =
/  y = ly/n =
                                          and
     n = number of calibration points.

The equation of the line fitted to the data is written as

     Y = y + b(x-x) = (y-bx) + bx = a + bx,

where Y =• predicted mean response for corresponding x,

      b = slope of the fitted line, and

      a = intercept where the line crosses the y-axis.
      b =
          Zxy -
                   n
                    ,2  .
                                               oo. o
                  n
           a = y - bx =
        . 00
Figure 2.3.  Calculation form for the method of least squares.

-------
                                               Section No.  2.3.2
                                               Revision No.  0
                                               Date  July 1,  1979
                                               Page  17 of 27

  set  of' points.  On  the  y-axis  of  the  graph,  locate and plot  the y
  intercept  (a).   Using  the  equation  y  =  a  +  bx,  calculate the
  predicted  y  value   using  the  80% URL  concentration  for the x
  value.   Plot this  second point  on the graph.   Draw a  straight
  line   through  these   two   points  to  give ' a   best-fit   line.
  Figure  2:2 shows a  calibration line plotted  using  this  procedure.
  Steps 9 and 10 should be repeated 'for  the NO values.
                                              X
      11.  After the best-fit  line has been  drawn  for the NO and
  the  N0x calibrations,  determine whether the analyzer  response is
  linear.   To  be considered  linear,  no  calibration  point should
  differ  from the best-fit line by more than  2%  of full scale.  A
  simple  test for linearity  can be made by plotting a point 2% of
  scale above and 2%  of scale below the point  where  the  fitted line-
  crosses the 0.4-ppm line.  Repeat this procedure where the fitted.
  line crosses  the 0.1-ppm line. .Draw a  straight line through the'
.  +2% points and the  -2% points  (Figure 2.2).   These  two lines de-
  fine the limits between which the calibration points can fall and
 the  calibration  curve  be  considered  linear.  Repeat  any points
 falling outside .these limits to eliminate  calibration errors; if
 the repeated  points still fall outside  the limits,  consult the
 manufacturer's manual on how to correct the nonlinearity.
 2 •4 •3   Gas Phase Titration Procedure  for NO,. -  Having completed
 the calibration of  the  NO  and NOx channels, the NO2 channel may
 now  be  calibrated  by  GPT.   The  Figure  2.1  (backside  of  the
 calibration data  form)  allows for  the  systematic recording of the
 data  determined during  the calibration of  the NO£  channel of the
 analyzer.   Do  not readjust  zero and span knob .settings for the NO
 and the  NO   channels.
           .X
      1.    The  N02 zero  adjustment  was made in  step 8-d  of the
 N0/N0x   calibration  and  need not  be  repeated.    Record the  de-
 termined response  under column 12 of the N02 calibration table.
      2.   N02   impurities,    [NO2lIMp,   found  in   the   reference
 standard must be recorded under column 10;  if there is  no impu-
 rity, disregard this column.

-------
                                             Section No. 2.3.2
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 18 of  27

     3.   Adjust  the  NO flow  (FNQ)  to generate an NO concentra-
tion  near 90%  of the  URL.   Dilution  air and  O3  generator air
flows* should be the- same as used in the calculation of  specified
conditions  in  Subsection  2.3.2.   Sample  this  NO  concentration
until the NO  and NO  responses stabilize.  Using the NO calibra-
                    *i
tion  relationship determined in step. 10,  measure and record the
NO  concentration under column  8,  [NO]~_.T-,.   Using the  NO  'cali-
                                       UKJ.LJ               x
bration relationship  obtained  in step 10, measure and record the
NO  concentration under column  7  [NO ].   Record both values  on
  X                                   X
the line marked  "ORIG."
     4.   Adjust  the  O3  generator to  produce  sufficient  O3  to
decrease  the  NO  concentration from  90%   to  10% of  full   scale.
This will be equivalent to 80% of the URL  of the uncalibrated' N02
channel.  The  decrease must not exceed 90%  of the NO concentra-
tion  determined in step 3 .  After  the analyzer responses  stabi-
lize,  determine  the new NO and NO  concentrations from  their re-
                                  x
spective calibration relationships .  Record the  NO  concentration
                                                  **
under column 7, and the NO concentration under column 9,       -..
                                                             Ktrl
The [NO]OR_G. will be the same value determined in step 3.
     5.   Calculate  the resulting NO2  concentration,   [NO-]    ,
using  Equation 2-11,  and  record  it  under column  11,   [NO2]QUT.
         = CNO]ORIG - [N°]REM +                  •   Equation  2-11
                                  NO    0    D
If  there  was no  NO2  impurity  in the  NO  reference  standard,
Equation 2-llA may be used to Calculate [NO-] Q _.

     [N02]OUT = -[NO]ORIG ~ [NO]REM'   '        .     Equation 2-llA
     6.   Adjust the  NO2 span  control  to obtain  a  recorder re-
sponse  using  Equation 2-10A.    This  equation  is derived   from
Equation 2-10  by substituting ' [NO_2]QUT  and ZNQ  for  [NO]QUT and
ZNO'

-------
                                              Section No. 2.3.2
                                              Revision No. 0
                                              Date July 1, 1979
                                              Page 19 of 27
                                 [NO ]
  Recorder response (% scale)  = (—*     x  100)  + ZXT^
                                   UKIj               NO_
                                                    Equation 2-10A
 Note:   If  the analyzer has  only one or two span controls,  the
 span adjustments are made on  the NO channel  or  on the NO and NO
 channels,  and  no further  adjustment  .is made here for  NO  .   If
 substantial  adjustment of  the NO2 span control  is necessary,  it
 may be  necessary  to . recheck  the zero  and  span  adjustments.
 Record the NO2 recorder response  under column 12.
     7. ••  While  maintaining   all  other  conditions,   adjust  the
 ozone  generator to obtain two  other concentrations of NO  evenly
 spaced between the 80% URL point and  the zero point.   Record the
 information  for  each point  on the respective  lines  of the  NO
 calibration  table.
     .8,   Repeat  steps  9,  10, and 11  of Subsection 2.4.2 for the
 N02 .recorder reading, column -12,  and the corresponding calculated
 concentration  [NO2"]QUT, column 11.                 .  -
 2•4•4  Example  NO and NO2 Calibration  -
     1.   Complete  steps  1  through 5 and  step  7  of the NO/NO
 calibration  before  starting the calibration,  to  document all  in-
 formation concerning  the  station, analyzer,  calibrator  reference
 standard,  and  person  performing  the   calibration.   Because
 analyzers have different  operating  characteristics,  consult  the.
 manufacturer's  operation  manual before starting the  calibration
 procedure.
    .2.   Select  the  operating range (ppm)  of the  analyzer to be
 calibrated.   For  this  example,  assume •that  - all  three  channels
 (NO, N0x,  and  N02)  will  be  calibrated on  the  range  of  0 to
 0.5-ppm.
     3.   Be sure that  the recorders  are operating properly and
are connected  to  the correct  output terminals  of  -the analyzer.
     4.   Connect the analyzer's  sample  line  to the  manifold of
the calibrator.

-------
                                              Section  No.  2.3.2
                                              Revision No.  0
                                              Date  July 1,  1979
                                              Page  20  of 27

     5.   Adjust  the diluent  air  flow (FD) and the  O3 generator
air  flow (FQ) to  obtain the  flows  close to those determined  in
Subsection  2.3.2.    Remeasure  these  flows  for each  calibration:
                      3
          FD = 2460  cm /min @  STP,  and
           i
           O
Fn = 260 cm /min @ STP.
Record the sum of.FD  and FQ under column  1,

          FD + FQ = 2720 cm3/min @ STP.

     6.-   Allow the analyzer to sample the zero  air until the NO,
NO  , and  NO0  responses stabilize.   Then  adjust  the analyzer zero
 : «?V        ^
control(s), and  offset the  analyzer zero adjustments  to  +5% of
scale  to  facilitate  observing any negative  zero drift.  Record
the  stable  zero  air  responses of 5%  under  column 4  for  NO and •
column 6  for NO .  Record the  NO., zero air response in  column 12'.
               x                z
     7.   Adjust  the  NO  flow from 'the,  standard NO  cylinder to
generate  an  NO concentration  of approximately  80% of the URL of
the  NO  channel;  and measure the NO  flow  (F   ),  and record under
column 2  on the 80% URL line.
                                3
                      F   = 22  cm /min.
     8.   Calculate  the  exact  NO.  concentration .[NO]QUT  using
Equation  2-8,


          [NO]OUT = 2720 +°22  = 0'405 ?**'
Record this value on the 80% URL line of  column  3.  Calculate the
exact N0x concentration [NOx]OUT, using Equation 2-9,

          rxio i    - 22 x (50.5 + 0)    _  . __
          CN°x]OUT	2720 + 22	   = °-405 PPm-
Record this value on the 80% URL line of  column  5.
     9.   Sample the generated concentration until the  NO and NO
                                                                X
responses  have  stabilized.   Calculate  the   expected recorder

-------
                                             Section No. 2.3.2
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 21 of 27

response  (%  scale)  for this concentration,  using Equation 10,

        Response (% scale) = (g'^05 x 100) + 5 = 86%.

Adjust the NO span.control to obtain a recorder reading of 86% of
scale, and record this  reading on the 80% URL  line  of column 4.
     Substitute  the  NO   values  in Equation 2-10; a  response of
                       X
86%  is  determined.   Adjust the NO   span  control  and record this
                                  X
reading under column 6.
     10.   Generate  two  approximately evenly  spaced  points  be-
tween zero and the 80% URL by changing either FNQ or F_.  In this
example,  F   was changed.  Allow  each trace to  stabilize before
moving to the next calibration point.  When each trace has stabi-
lized, record the required data  in the appropriate column of the
NO/NO  calibration table.
     Jt
     11.  Plot  the  analyzer response  in %  chart from  column  4
(y-axis)  versus  the  corresponding  calculated  concentration
[NO]QUT from  column  3  (x-axis.).   A straight line  of  best fit is
now calculated by the method of least squares.   For this example,
the slope (b) is 200 with a y-intercept of 5%.   Plot the calibra-
tion relationship as in Figure 2.2.  To check linearity, draw the
+2%  and  -2%  lines  parallel to  this calibration  line.   In this.
example,  the analyzer response is linear.   '  •                  •
     12.  Adjust the NO  flow (FNQ) to generate  an NO concentra-
tion near 90% of the NO range.   For this  example,  a concentration
   90
of TOO x °'50 ppm (or °'45 ppm) is used-   Tnis value need only be
approximated.   After, the  response  has   stabilized,  use  the NO
calibration relationship  determined' in step 11 to arrive at the
actual  concentrations.   Record  this  value  oh  the'  line marked
"ORIG" under  column  8  [NO]ORIQ.   If the  NOx   channel is moni-
tored,  determine NO  . concentrations  from  the  NO   calibration
                    x         '                     X
relationship, and record the values under column 7.
     13.   Adjust the  O3   generator  to produce  sufficient.  0_  to
reduce the NO concentration  from  90% of  full scale (0.45 ppm) to
    of full scale (0.05  ppm).  Determine  the actual NO concentra-

-------
                                              Section No.  2.3.2
                                              Revision No.  0
                                              Date July 1,  1979
                                              Page 22 of 27

           '      .                                       '             *
tion  remaining from the calibration relationship,  and record the
value  on the 80% URL  line .under column 9  [NO]...,,....    [NO]^nTO is
                                               Khjl"]         UKlLr
the same value  determined in  step  12.
     14.   The   resulting  NO2  concentration  is  now  calculated.
Because  there were no NO2 impurities  present in the  NO  standard
cylinder,  Equation  2-11A may  be  used.

           [NO2]QUT  = 0.45 - 0.05 = 0.40 ppm NO2.

     15.   Calculate  the   required   recorder   response   using
Equation 2-10,

   Recorder response (% scale) = '(—-^ x 100)  + 5  =  85%.

Adjust the NO2  span control to obtain a recorder  response  of 85%.
Record this value under column 12.
    ..16.  Adjust the  ozone  generator  for two additional  concen-
trations of NO2/- approximately evenly spaced between  the  80% URL
point  and  the  zero point.    Record  the required data under  the
appropriate columns of  the  NO2 calibration  table  (Figure  2.1).
     17.  Plot  the  analyzer  response  in  % chart from column  12
(y-axis)   versus   the  corresponding   calculated   concentration
t^a-'oUT  from  column  11  (x-axis).   Proceed as  in  step  11  to
determine a straight line of best fit and linearity.'

-------
                                     Section No. 2-3.2
                                     Revision No. 0
                                     Date July 1, 1979
                                     Page 23 of 27
Calibration
point
Zero
set point
80% URL
1
2


1
tN°2W
(x)
o.o
0. 40
J. 2.0
0. /O.


2
tN°xWG
^. ^r

-------
   0.5
   0.4
1.  0.3
n.
Q.
o
o
   0.2
   0.1
0.1
                                                   Section No. 2.3.2

                                                   Revision No. 0

                                                   Date July 1, 1979

                                                   Page 24 of 27
               Efficiency- = Slope x
                          0.2
                    0.3
                          [N°2JOUT' Ppm
         Figure  2.5.  Converter  efficiency relationship

-------
                                              Section No.  2.3.2
                                              Revision No.  0
                                              Date  July  1,  1979
                                              Page  25 of 27

 2.5  Determination of NO^ to NO Converter Efficiency
     A  data form (Figure 2.4)  has been developed  for determining
 the  converter efficiency.   The  following  procedure is  for  use
 with this data form.
     1.   Values  for columns 1,  2,  and  3  of the  converter  effi-
 ciency  data  form  are  taken directly, from  the NO2  table of  the
 calibration data form.
     •a.   Column  1   [NO2]OUT is  from column  11  [NO2].0UT of  the
 table.                 .
     b.   Column  2   [N0x] ORIG is  from the  line marked "ORIG"  of
 column  7  [NO  ]  of the table; this value will be the same  for  all-
            •«*• '                    .       '
 lines of column 2.
     c.   Column  3   tN°x3REM is  from the appropriate calibration
 points of column 7 of the table.
     2.   Calculate  -the  quantity  of NO2  converted to NO,  labeled
            for each  point using Equation 2-12.

 ,  [N°2]CONV = [N02]OUT * ([NOx]ORIG ~ [N°x3REM)-     Equation 2-12

     3.   Plot  [N02]CONV (y-axis) versus  [NO2]QUT  (x-axis),  the
 converter efficiency curve  (Figure 2.5),  and calculate the slope
 (b) of  the  curve using  either  an  appropriate  calculator or  the
 calculation form  (Figure 2.3)  for the  method  of  least squares.
     4.   Multiply the slope (b) of the curve by 100 to determine
 average converter  efficiency;  if  the efficiency  is  <96%, either
 replace or service the converter.
 2 . 6  Calibration Frequency
     To ensure accurate measurements of the NO and NO- concentra-
 tions,  calibrate  the analyzer  at  the time  of  installation,   and
 recalibrate  it:
     1.    no later than  three months after  the most recent cali-
bration or performance audit which indicated analyzer response  to
be  acceptable; or
     2.    following any one of  the activities listed below:

-------
                                              Section No. 2.3.2
                                              Revision No. 0
                                              Date July 1, 1979
                                              Page 26 of 27

          a.   an  interruption of more  than a few  days  in ana-
lyzer operation;
          b.   any  repairs which  might affect  its  calibration;
          c.   physical relocation of the analyzer; or
          d.   any  other  indication  (including excessive  zero or
span drift)  of possible  significant  inaccuracy of the analyzer.
Following any  of  the" activities listed in item 2 above,  the zero
and  span should  be  checked  to- determine  if" a  calibration  .is
necessary.   If the analyzer  zero  and span  drifts  do  not exceed
the calibration limits  in Table 9.1 of Section 2.0.9,  Subsection
9.1.3,  a  calibration need-not be performed.   If  either  the zero
or span drift exceed their respective calibration limit,  investi-
gate the cause of the drift, take corrective action and calibrate
the analyzer.

-------
                                                       Section No.  2.3.2
                                                       Revision No.  0
                                                       Date  July  1,  1979
                                                       Page  27 of 27
      Table  2.1.   ACTIVITY MATRIX  FOR CALIBRATION PROCEDURES
Calibration
Activities
Acceptance limits
                     Frequency and method
                        of Measurement
                       Action if
                       requirements
                       are not met
Calibration
 gases
Sec 2.0.7, Subsec 7.1-
                    Assayed against an
                    NBS-SRM quarterly
                    Sec 2.0.7
                       Working  gas
                       standard is
                       unstable and/
                       or measurement
                       method is  out
                       of control;
                       take  corrective
                       action such as
                       obtaining  new
                       calibration gas
Dilution gas
Zero air,  free of
contaminants;
and Sec 2.0.7,
Subsec 7.1
                    See TAD
                                                2
                       Return  to  sup-
                       plier,  or  take
                       appropriate cor-
                       rective action
                       with  generation
                       system
Multipoint
 calibration
 (GPT)
1.
t_, < 2 min
 K —
PR > 2.75  ppm-min
Method
.1.  Subsec 2.3.2
                 2.  Use calibration
                 procedure in Subsec
                 2.4; also TAD  and
                 the Federal Register
                3.  Converter Effi-
                ciency > 96%
                        2.  Subsec 2.4,
                        TAD ,  Federal
                        Register  and
                        Fig 2.1; see Sub-
                        sec 2. 1 for fre-
                        quency

                        3.  Subsec 2.5 and
                        and Figs 2.4 and
                        2.5
1.   Adjust flow
conditions and/
or reaction
chamber;  volume
to meet sug-
gested limits

2.   Repeat the  '•
calibration ..
                                           3.   Replace or
                                           service the
                                           converter

-------

-------
                                              Section No.  2.3.3
                                              Revision No.  0
                                              Date July 1,  1979
                                              Page 1 of 10

 3.0  OPERATION AND PROCEDURE
      Essential  to quality  assurance  are scheduled  checks  for
 verifying the  operational  status  of the monitoring  system.   The
 operator should -visit  the  site at  least once each  week.   Every
 two weeks  a Level  1  zero  and span  check must  be  made  on  the
 analyzer.'  Level 2 zero and span  checks  should be conducted at a
 frequency desired  by the user.
      In  addition .an independent precision .check  between 0.08  and
 0.10  ppm must  be   carried  out  at  least once every two  weeks.
 Table 3.1  at  the   end  of  this section  summarizes  the   quality
 assurance  activities  for  .routine  operations.'   A discussion  of
 each  activity  appears in the following  sections.
      To  provide for documentation  and accountability of  activi-
 ties,  a  checklist  should be compiled  and then filled out  by  the
 field operator as  each activity is  completed.  An example  check-.
 list  is  Figure  3.1.
      In  Subsections 3.1 and 3.2,  reference is made  to the sam-
 pling  shelter  and  the sample inlet .system.   The  design and con-
 struction  of  these components  of  the  sampling   system" are  not
 within the scope of this document.   An in-depth study of'these  is
 reported • in  Field Operations Guide for Automatic Air Monitoring
 Equipment,  Publication  No.  APTD-0736,  PB 202-249  and PB  204-650
 U.S.  Environmental.'Protection  Agency,  Office of  Air Programs,
 October 1972.
 3.1  Shelter
     The  shelter's role  in quality assurance is to  provide a
 temperature-controlled environment' in which the  sampling   equip-
ment  can  operate- at optimum performance.   The mean shelter tem-
perature  should be between  22°  and 28°C  (72°   and 82°F).  • A
thermograph  should  be  installed at  the shelter  to  record daily
fluctuations in temperature continuously.  Fluctuations  greater
than  ±2°C (±4°F)  may  cause  the  electronic   components  of  the
analyzer to  drift and may introduce error into the data;  thus the
fluctuations outside of the  specifications  should be identified,

-------
                                             Section No. 2.3.3
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 2 of 10
Site
Site
Site
      ID
        OOI
     location

     address
          Inspect thermograph for temperature variations greater
          than '±2°C  (4°F).  Identify time frame of any tempera-
          ture level out of tolerance
          Comments
          Inspect sample introduction system for moisture, parti-
          culate buildup, foreign objects, bre§kage, leaks
          Comments

          Is sample 'line connected to manifold?
          Comments
          Inspect data recording system
                                             OK
                                           Corrective
                                          action taken
             Legibility of trace
             Ink supply
             Paper supply
             Chart speed selector
             Signal range switch
          0  Time synchronization

          Comments
o

o

o

o

o
          Inspect analyzer operational parameters
                                             OK
                                           Corrective
                                          action taken
             Sample flow rate
             Oven temperature light
              flashing
             Analyzer in sample mode
             Zero and span potentiom-
              eters locked at correct
              setting
          Comments
Figure 3.1.  Example of an operational checklist (front side).

-------
Signature
                                              Section No.  2.3.3
                                              Revision No.  0
                                              Date July 1,  1979
                                              Page 3  of 10
       6 .  Zero the  analyzer
  *    7.   Is unadjusted  zero within  tolerance?
            Comments    /je^^y-   0£/    0>£  »: S/..
   i/   8.  Span the  analyzer.

   ^   9.  Is unadjusted  span within  tolerance?
           Comments
  v   10.  Enter. zero and span values  on  span  check data  form

  vX  ll.  Return to sample mode

  \/  12.  Record cylinder pressure of zero  and  span tanks

               Zero air       /3OO

               Span air       /£
      13 .  Close valve on zero and span tanks
Figure 3.1.  Example of an operational checklist (backside).

-------
                                              Section No.  2.3.3
                                              Revision No.  0
                                              Date  July  1,  1979
                                              Page  4 of.  10
 and  the data  for  the affected  time  period should be' flagged to
 indicate possible  discrepancies.
 3.2   Sample  Introduction  System
      The  sample introduction  system  consists of an intake port,
 the  particulate and  moisture traps,  the  sampling  manifold and
 blower, and  the sampling line to  the  analyzer.  The field opera-
 tor,  as part  of  the  quality assurance  program,  should inspect
 each  of  these  components  for  breakage,  leaks,  and  buildup of
 particulate  matter or other  foreign  objects; check for moisture
 deposition  in  the sample  line or manifold;  see  that the sample
 line  is connected  to the manifold; .see that  any component of the
 sample  introduction system that  is not within tolerance is either
 cleaned or  replaced  immediately.   See  Section 2.0.2  for  more
 details.
 3.3   Recorder
      During  each weekly visit to  the  monitoring  site, the field
 operator should use  the  following list to  check the recorder for
 proper  operation:
      1.   Ink trace for visibility.
      2.   Ink level in reservoir.
      3.   Chart paper for supply.
      4.   Chart speed control setting.
      5.   Signal input range switch.
      6.   Time synchronization.
 Any operational parameter that  is  riot within  tolerance  must be
 corrected immediately.
 3.4  Analyzer
      Specific instructions in the manufacturer's manual should be
 read  thoroughly before attempting  to operate  the  analyzer.   As
 part  of  the  quality assurance program,  each site  visitation
 should  include  a visual inspection of the  external  operation of
 the analyzer, the  zero  and span checks, and  a biweekly precision
point check.

-------
                                             Section No. 2.3.3
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 5 of 10

3.4.1   Visual Inspection  -  During the  visual  inspection,  the
field  operator should inspect  the  external operating parameters
of  the  instrument.   The  parameters of  concern will  vary from
instrument  to  instrument,  but  in general they  will  include the
following:
     1.   .Correct setting of  flow meters and regulators;
     2.   Cycling of temperature control indicators.
     3.   Temperature level if equipped with a pyrometer.
     4.   Verification that the  analyzer  is in the sampling mode
rather than the zero or the calibration mode.
     5.   Zero and'span potentiometers locked  and  set at proper
values.
3.4.2  Zero and Span Checks  - Zero  and span checks  must, be used
to document within-control conditions.  The purpose is to provide
interim checks on the response of the instrument to known concen-
trations.  If  a response  falls  outside of the prescribed limits,
.the analyzer is considered  out  of control, and '.the, cause must be
determined and corrected.   A quality control chart can be used to
provide  a visual check  to determine  if  the analyzer  is  within
control conditions.   A zero check should be conducted at the same
time that the span check is performed.
     A system  of Level  1  and Level  2 zero  span checks  is  recom-
mended.   These checks  must be  conducted  in accordance  with the
specific  guidance   given  in  Subsection  9.1 of Section  2.0.9.
Level  1  zero and  span checks must be conducted every  two  weeks.
Level 2  checks should be  conducted in between the Level 1  checks
at a frequency desired by the user.   Span concentrations for both
levels should  be between  70 and  90% of the -measurement  range.
The data  should be  recorded on  the  zero span check  form,  Figure
3.2.
     Zero and span data are to be used to:
     1.   provide data to allow  analyzer  adjustment  for zero and
span drift;
     2.   provide  a decision point on  when   to  calibrate  the
analyzer;

-------
                                             Section No.  2.3.3
                                             Revision No.  0
                                             Date July 1,  1979
                                             Page 6 of 10
Site ID
Location
                                      Pollutant

                                      Analyzer
Address  33S&
                                      Serial number

Date
J-/-77
J-/r-77
5-2.9-77





























^ /
Technician
<^:.X
^r-. T.
^l.T.

•




























Unadjusted
zero,
% chart '
d:o
^r. £
.r. o












•

















Span
concentration,
ppm
0. 4o
a. 
-------
                                             Section No. 2.3.3
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 7 of 10

     3.   provide a  decision  point on invalidation of monitoring
data.
Items  1 and  2  are  described in  detail  in Subsection  9.1.3 of
Section  2.0.9.   Item 3  is  described in  Subsection  9.1.4 of the
same section.
     When  the  response from a span  check is  outside the control
limits, the cause for the extreme drift should be determined, and
corrective action should be taken.   Some of the causes for drift
are listed below:
     1.   Lack of preventive maintenance.
     2.   Fluctuations in electrical power supply.
     3.   Fluctuations in flow.
     4.   Change in zero air source.
     5. '  Change in span gas concentration.
     6.   Degradation of photomultiplier•tube.
     1.   Electronic and physical  components  not within manufac-
turer's specifications.                            •
Corrective actions  for  the above can  be found  in  the  manufac-
turer's instruction/operations manual.
3.4.3   Precision Check  -  For continuous analyzers,  a  periodic
check  is  used  to  assess the  data  for precision.  A one-point
precision  check  must be  carried  out at least  once  every 2 'weeks
on each  analyzer at an  NO2 concentration- between  0.08  and 0.10
ppm.   The  analyzer must  be  operated in its normal sampling mode,
and the precision test gas  must  pass through all filters, scrub-
bers,   conditioners,   and other  components  used  during  normal
ambient sampling.  The standards  from which precision check test
concentrations are obtained must be  traceable  to NBS-SRM.  Those
standards used for calibration or auditing may be used.
     Precision Check Procedure
     1.   Connect the  analyzer  to  a  precision  gas  that  has  a
concentration between  0.08  and 0.10  ppm.   An NO-  precision gas
may be generated by  either  the GPT or a NO2 permeation tube.  If

-------
** -»t be made
                Section No.  2 3
                     i0n NO.  6
conjunction
             Precision.

-------
                                                  Section No.  2.3.3
                                                  Revision No.  0
                                                  Date July  1,  1979
                                                  Page 9 of  10
U)
3

s-i
o
-p
ra
M
u
\?

 •4
           03
u
G
(U
M e
i 0 S
•H cu a

S-I 0) U
CM -P G
O
U .
(1)
N
C


G
Station
locatio

-P
(C
Q
G
O >i
-H jQ
cn
•H .X
u u
^^
E
\
^^
V


"
TJ
_ •
\J




.
V
Q
^

o
1

&
^"
*C\
^





^

•
M



5
•^
Kl
^
V)

1
1
^
\
\
\


.*
'O
t
^J




.
*J
Q
^

^


^
^
•
^





£
^
•
.^



-3
M
^
w
^
*

i
r-
V
*^
^


k^
^
•
\J


































































































































































































































































































































































































































































































































































































• \






























_































































                                                                !-)
                                                                O
                                                                U
                                                                1)

                                                                "u

                                                                G
                                                                O
                                                                •H
                                                                W
                                                                •H
                                                                O
                                                                0) .
                                                                S-I
                                                                a
                                                                g
                                                                w
                                                                S-l

-------
                                                        Section No. 2.3.3
                                                        Revision  No.  0
                                                        Date July 1,  1979
                                                        Page 10 of 10
                   Table  3.1.   DAILY ACTIVITY MATRIX
Characteristic
Shelter temper-
 ature
Sample in.tro-
 duction system
Recorder
Analyzer oper-
 ational set-
 tings
Analyzer oper-
 ational check'
Precision
 check
Acceptance limits
Mean temperature be-
tween 22° and 28°C
(72° and 82°F),  daily
fluctuations not
greater than ±2°C (4°F)
No moisture, foreign
material, leaks, ob-
structions; sample line
connected to manifold
1.  Adequate ink sup-
ply and chart paper
2.  Legible ink traces
3.  Correct settings of
chart speed and range
switches
4.  Correct time
1.  Flow and regulator
indicators at proper
settings•
2.  Temperature indi-
cators cycling or at
proper levels
3.  Analyzer set in
sample mode
4.  Zero and span con-
trols locked
Zero and span within
tolerance limits as
described in Subsec
9.1.3 of Sec 2.0.9
Assess precision as
described in Sec 2.0.8
and Subsec 3.4.3
 Frequency and method
    of measurement
 Edit thermograph
_ chart daily for
 variations greater
 than ±2°C (4°F.)
 Weekly visual inspec-
 tion
 Weekly visual inspec-
 tion
 Weekly visual inspec-
 tion
 Level 1 zero and span
 every 2 weeks; Level
 2 between Level 1
 checks at frequency
 desired by user
 Every 2 weeks,
 Subsec 3.4.3
Action if
requirements
are not met
1.  Mark strip
chart for the
affected time
period
2.  Repair or
adjust tempera-
ture control
system
Clean, repair,
or replace as
needed
1.  Replenish
ink and chart
paper supply
2. • Adjust re-
corder time to
agree 'with clock;
note on chart
Adjust or repair
as needed
1.  Isolate
source of error,
and repair
2.  After cor-
rective action,
recalibrate
analyzer
Calculate, re-
port precision,
Sec 2.0.8

-------
                                             Section No. 2.3.4
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 1 of 5
4.0  DATA REDUCTION, VALIDATION, AND REPORTING
     Quality assurance activities for data reduction, validation,
and reporting are summarized in Table 4.1 at the end of this sec-
tion.
4.1  Data Validation
     Monitoring  data  of poor quality may be  worse than -no data.
Data validation 'is one activity of a quality assurance program to
screen data for possible errors or anomalies.  Reference 8 recom-
mends  several  statistical  screening procedures  for  ambient air
quality data that  should be applied to identify gross data anom-
alies.  Subsections 4.1.1 and 4.1.2 recommend two data validation
checks.
4.1.1   Span Check Drift  -  The  first  level  of  data  validation
should be to accept or reject monitoring data based upon routine
periodic  analyzer  checks.   It  is -recommended that  results  from
the Level 1 span checks discussed in Section 2.3.3.be used as the
first level of data validation for accepting data.  This means up
to two  weeks  of monitoring data may be  invalidated  if the  span
drift  for a Level  1 span check is equal  to  or greater than 25%.
For this  reason,  it may  be desirable  to perform Level 1 checks
more often  than the minimum  recommended frequency of  every two
weeks.
4.1.2  Edit of Strip Chart
     The.  strip  chart  should  be  edited to  detect signs  of the
monitoring  system's  malfunctions  that  result in traces  on the
chart that do not  represent "real" data.  When reviewing a strip
chart,  typical  indicators of malfunctions to watch for are:
     1.   A straight  trace  (other  than minimum  detectable) for
several hours.
     2.   Excessive noise  indicated by a vide  solid  trace,  or
erratic behavior such as spikes that are sharper than is possible
with  the  normal  instrument  response   time.   Noisy outputs  may
occur when analyzers are exposed to vibrations.
     3.   A long steady increase or decrease in deflection.

-------
                                              Section No. 2.3.4
                                              Revision No. 0
                                              Date July 1, 1979
                                              Page 2 of 5

      4.   A  cyclic  trace  pattern  with  a  definite time  period
 indicating a sensitivity to  changes in temperature or parameters
 other than NO2  concentration.
      5.   A  trace  below the  zero  baseline  that may  indicate  a
 larger than  normal  drop in  ambient room  temperature  or  power
 line voltage.
      6.   Span  drift equal to or  greater  than 25%,  Subsec.  9.1.4
 in Section 2.0.9.
      Void data for  any  time  interval for which a malfunction  of
 the sampling  system  is detected.
 4.2  Data Reduction
      To obtain hourly average  concentrations from a strip  chart
 record, the following procedure may  be  used.
      1.'   Be  sure  the strip chart record  for the sampling period
 has a zero trace at  the  beginning  and end  of  the sampling period.
      2.    Fill  in  the identification data called'for at the top
 of the hourly average data form, Figure 4.1.
      3.    Use a  straight edge to draw  a line  from the  zero  base-
 line  at the start  of the sampling period to  the zero baseline at
 the end of the  sampling period.   This line  represents  the zero
 baseline  to be used  for  the sampling period.
      4.    Read the zero  baseline in  % of chart at the midpoint of
 each  hour  interval,  and  record the value on the data form.
      5.    Determine  the  hourly averages,  for  the  interval  of
 interest  between two vertical hour lines,  by placing a transpar-
 ent straight  edge parallel to .the horizontal chart division  lines '
 and by adjusting the straight edge between the lowest and highest
points  of  the trace  in  that interval so that  the  area above the
 straight  edge and bounded  by the  trace  and the hour lines  is
estimated  to  equal  the  area below the  straight  edge and bounded
by  the trace and'hour lines,'as shown below.

-------
                                              Section-No. 2.3.4
                                              Revision No. 0
                                              Date July 1, 1979
                                              Page 3 of 5.
Llty Z34/TZ7A/ , nff/O
Site location -2,100 TVA/S-^ / *•*/?
L*flG CKc 3T y — "7V7// A/^f^A/'
CALIBRATION CURVE: Slope (b) = ^
Site number 3^
Pollutant ,
&






















Check






















—

Difference
Ori;
/5
j-j





















— ' I..-
••^"•HMBM^M
Check























"
- —i- i •,
y, Add + 5
Ori
2&
•Z-"2~






















Check

























x, ppm
Orig
0. Og
























Check
























Figure 4.1.   Sample  data   form  for  recording  hourly  averages.

-------
                                              Section No.  2.3.4
                                              Revision No.  0
                                              Date July  1,  1979
                                              Page 4 of  5

 Read and record the  percentage of chart deflection on  the hourly
 average data form.   Repeat the procedure for all the hour inter-
 vals sampled  which  have  -not been  marked  invalid.   Record all
 values  on  the  hourly  average data form  in  the  column headed
 "Reading -  Original".
      6.   Subtract the zero baseline value from the reading value
 and record  the difference on the hourly averages form.
      7.   Add  the  percentage  ,of  zero  offset,   +5,  to  the  dif-
 ference on  the hourly'averages form.
      8.   Convert  the  % chart values  to  N02   concentrations  in
 ppm using the most recent  calibration curve.   Record the ppm NO7
 values  in the  last column of the hourly averages form.
      An alternative  method for converting  %  chart to  ppm is to
 use the following' equation and thus  to eliminate steps 6, 7, and
 8 of the  above procedure.

                             y - y  .
                      PPm = slope  /           ,       Equation 4-1
where
          y  = recorder reading in % scale,  from  step  5,
        YZ  = zero baseline  in % scale,  from  step 4,  and
     Slope  = slope  of  the  calibration  relationship,   as  deter-
             mined in Section 2.3.2.
4.3  Data Reporting
     Transcribe  information  and  data  from the hourly averages
form to. a SAROAD hourly data  form; see  Section  2.0.3  for details
and  basic instructions for filling out  the  SAROAD.   If the  data
are  to  be  input to  the National  Aerometric Data  Bank,  obtain
further  instructions  from  the  SAROAD  User's  Manual,  APTD-0663.
                                  .AREA ABOVE LINE
                 STRAIGHT EDGE ^V^V    BE1-°H LINE
                  1200	1300	1400      T500	f600(

-------
                                                        Section No.  2.3.4
                                                        Revision No.  0
                                                        Date  July  1,  1979
                                                        Page  5 of  5
           Table 4.1.  ACTIVITY MATRIX FOR  DATA  REDUCTION
Activity
Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Data validation
 Span check
 • drift
Level 1 span check
<25%, Sec 2.3.3
Perform Level 1 check
at least every two
weeks, Sec 2.3.3
Invalidate data;
take corrective
action; increase
frequency of Level
1 checks until
data are acceptable
 Edit strip
  chart
No signs of malfunc-
tions
Visual.ly edit each,
strip chart;  see
Subsec 4:1
Void data for
time, interval
for which
malfunction of
sampling system
is detected
Data reduction
Stepwise procedure
for data reduction
(Subsec 4.2)
For each strip  chart,
follow the method
given, in Subsec 4.2
                                                               Review the
                                                               reduction
                                                               procedure
Data reporting
Follow Reference  6  in
transcribing data to
SAROAD hourly data
form
Visually check
Review the
data trans-
cription
procedure

-------

-------
                                              Section No.  2.3.5
                                              Revision No.  0
                                              Date July 1,  1979
                                              Page 1 of 2
 5.0   MAINTENANCE
 5.1   Preventive Maintenance
      Because  maintenance  requirements vary  from  instrument • to
 instrument,  the  supervisor   should  refer • to  the manufacturer's
 manual  for a  specific  analyzer.   After  becoming  familiar  with
 these maintenance requirements, the  supervisor should develop  a
 suitable preventive maintenance schedule.
 5.2   Corrective Maintenance
   .   Corrective  maintenance  is  defined  as  nonscheduled  main-
 tenance  activities  that become .necessary due  to system malfunc-
 tions.  A  few  examples of corrective maintenance are:  replacing
 a  damaged  pump diaphragm,  cleaning  a clogged  sample  line,  and
 replacing  a  N02  converter.   The  need for corrective maintenance
becomes  apparent  'as  the 'operator performs  the operations  de-
scribed  in Section  2.3.3.,'  When the  need for  corrective mainte-
nance arises,  the operator should refer to the owner's manual for
trouble-shooting  procedures.   A detailed maintenance  record of
corrective activities should be kept on file for each analyzer at
the site to identify reoccurring malfunctions.   A maintenance log
appears in Figure  5.1.

-------
                                                Section No. 2.3.5
                                                Revision No. 0
                                                Date July 1, 1979
                                                Page 2 of 2
Site number
Sit? 1 ocat.ion

Site address
                              Pollutant

                              Instrument
                              Serial number
Date
 Tech-
nician
   Event
initiating
maintenance
Maintenance
  activity
Comments
/-SB- 77
 c j;
                                                                TO
                                                    OF ±
                                                AT
                                                         J~-/O
                                                          TO Sz
         c JT
                              AT
                                                             rO JSS.
                                                          &V
                Figure  5.1  Sample maintenance log.

-------
                                           Section 2.3.6
                                           Date June 5,  1984
                                           Page 1
 6.0  AUDITING PROCEDURE
     An  audit is an  independent assessment of  the accuracy of
 data.   Independence is achieved by  haying  the audit made by an
 operator  other  than the  one conducting the routine field mea-
 surements  and by using  audit standards and equipment different
 from those routinely  used in monitoring.  The audit should be a
 true  assessment  of the  measurement  process  under  normal  op-
 erations  without any  special preparation or  adjustment of the
 system.   Routine quality control checks.(such as zero and span
 checks in  Section 2.3.3) conducted by the operator  are necessary
 for obtaining and reporting good quality data, but they are not
 considered part of the auditing procedure.
     Three audits are recommended:   two performance audits and'a
 systems audit.   The performance audits are described in Subsec-
 tions  6.1  and 6.2,   and the  systems audit  is  described in Sub-
 section 6.3.   These audits  are  summarized in Table  6.1 at the
 end of this section.
     Proper implementation  of an auditing  program will serve a
 twofold purpose:   (1)  to ensure  the integrity  of the data and
 (2) to assess .the data for accuracy.  The technique for estima-
 ting the  accuracy of  the data is given in Section  2.0.8 of this
 volume of  the Handbook.

 6.1  Performance Audit
     A performance audit  consists of challenging the continuous
 analyzer with known concentrations of NO- within the measurement
 range of the analyzer.  The difference between the  known concen-
 tration and the  analyzer response is  obtained,  and an estimate
 of the analyzer's accuracy is determined.
 6.1.1  Equipment - Personnel, equipment, and reference materials
used in conducting  audits must  be  independent  from  those nor-
mally used in calibrations and in span checks.

-------
                                           Section 2.3.6
                                           Date June 5, 1984
                                           Page 2

      Known  concentrations  of NO2 can be generated by the GPT of
 NO with 03 to  produce NO2 or  by the use  of an NO2 permeation
 tube and a  dynamic dilution system.  The NO and NO  channels are
 to be audited by dilution of the NO standard with zero air.  All
 audit gas  standards  must be traceable to NBS-SRM's as described
 by the  protocol  in  Sections 2.3.2  and 2.0.7.  All  flow rates
 should  be  measured using  a  calibrated soap bubble  meter or an
 eguivalently accurate procedure.
      Procedures  used   to  generate  NO,   NOX/   and  NO,  con-
 centrations, although not identical,  are somewhat similar to the
 procedures  described  in  Section  2.3.2.    If  during a  regular
 field audit,  the differences  recorded for most analyzers  are
 either  negatively  or  positively  biased,  a check of the' cali-
 brator  used in  routine calibrations  of the  analyzers  may  be
 advisable.
 6'1-2  Audit Schedule  -  The recommended audit schedule  depends
 upon  the purpose  for which the  monitoring  data are  being  col-
 lected.   For example,  Reference  3  requires that each  analyzer in
 State and Local  Air  Monitoring Networks  (SLAMS.)  be  audited  at
 least once per year.   Each agency must audit 25% of  the  refer-
 ence . or  equivalent  analyzers each quarter.   If  an  agency  op-
 erates less  than four reference  or equivalent  analyzers,  it  must
 randomly select  analyzers  for reauditing so  that one analyzer
 will  be  audited each calendar quarter  and each analyzer will be
 audited  at  least  once  a year.
      Reference  7  requires that each PSD reference or  equivalent
 analyzer be audited  at least  once a sampling quarter.  Results
 of these audits are used to estimate the accuracy of ambient air
 data.
 6'1-3  Audit Procedures for NO. NO^.  anj__gr^ _ Audits should be
 conducted by challenging the  analyzer  with  at  least  one audit
 gas of  known  concentration from  each  of the  following  ranges
within  the  measurement range  of  the  analyzer being audited:

-------
                                  Concentration Range,  ppm
           Audit Point                      NO
                1                      0.03  to  0.08
                2                      0.15  to  0.20
                3                      0.35  to  0.45
                4                      0.80  to  0.90

The  differences in  concentrations (ppm) between the  audit values
and  the measured analyzer values  are used to calculate accuracy,.
as described  in Section 2.0.8.
      Information on the  station,  analyzer,  audit device, refer-
ence materials,  and  audit  procedures  is  of prime  importance
since  the validity  of audit results  depends  on accurate docu-.
mentation.  The  following  procedures  and  ' audit report forms
(Figures  6.1,  6.2,  6.3,  and 6.4)  have been developed to aid in
conducting the  audit..
     Procedure  for NO  Audit -
     1.    Record the  station  name,  address,  analyzer manufac-
turer-,  model,  and  serial  number on  the audit summary  report.
     2.    Identify  the person(s)  performing  the audit and the
date that  the audit  is performed.
     3.    Record the type  of  audit  device  used.  If the audit
device was purchased,  record the  manufacturer's name,  model, and
serial number.   If the audit  device was  assembled by the user,
assign it. an identification number  so that audits  can be ref-
erenced to that  particular apparatus.
     4.    Identify  the  NO   cylinder   and  the  NBS-SRM used  to
verify the audit concentration.  Reanalyze the NO cylinder every
3 mo,  following the protocol  in  Section  2.0.7.  Note:   Section
2.0.7 recommends reanalysis every 6 mo; however, since quarterly
audits are recommended for continuous NO, analyzers3'7 and since
the  integrity of the audit  results must not be subject to chal-
lenge,  quarterly reanalyses.  of  NO  cylinders  are  recommended.
     5.    Identify  the device used  to measure  the  flow  rates.
     6.   Attach  appropriate pressure  regulator  to  the NO ref-
erence cylinder,  and take  care to flush  the pressure regulator

-------
                                            Section 2.3.6
                                            Date  June 5,  1984
                                            Page  4
1.   Station
                                       Analyzer
                   . v\
2.   Audit performed by

3.   Audit device  used
                                           Date
S-Vs-'VN.
4.   NO standard used
                                 ^. yj.\n* 2. Concentration
     Verified against NBS-SRM
                          By  ^s\v\
                                               Date
5.

6.
     Flow measured with
     Analyzer response  to  zero air  NO zero  =   g-. 0   %  scale
                                     N02 zero =   *r. a   %  scale
                                     NO  zero =   <=?•. o  %  scale
                        AUDIT SUMMARY
Analyzer
channel
NO
N02
N0x
Audit value,
ppm
0 .000
o . ^**
n. 3oo
o.oso
O . ooo
O.A^.V
a ^rvz
o. oKn




Response,
% scale
!T.£>
V».o
H^.«r
\5.0
^-.^
*0.£.
Kai.3-
\^.1?




Response,
ppm
o .000 .
OM«
a . Sot
o .oso
Q .OOO
o .ans*
0 . l«6,
o .oH^




Percent
difference
(Equation 4)
__-
-V \.V
+• v.s-
o.o

4-aA
-v M ,«r
-V -H. S




              Figure 6.1  Example of an audit summary form.

-------
Equation 1
PPM
        Y-a
         b
Equation 2
r  _ V	
            or F
                NO
                   _ V
                      avg
  Equation 3
[NO].
          [NO]
             STD
                    NO
                                                     Section 2.3.6
                                                     Date  June  5, 1984
                                                     Page  5
                                          Y " % scale
                                          b = slope  of the calibration line
                                          a = intercept of the calibration
                                              line
                                            FT = total flow rate, cm /min
                                           FNQ = flow rate of NOSTD> cm3/min
                                             V « volume measured with soap
                                                 bubble meter
   Equation 4

   X difference
                 [NO]R -  [NO]A
                                          "avg
                                                average time in minutes
                     [H01
                               x  100
  For N02, substitute  [N02]R and (NO]A
  For NOX, substitute  [NOX)R and tNOxJA
  Ecruation 5
  Equation 6
  "avg
Zero air:
                                        INO]__D •  concentration of NO -
                                                 standard used, ppm
                                         [NO]A « NO audit concentration,  ppm
                                         [NO]R * analyzer NO response, ppm
                                        [NO,]. * NO- audit conform GPT of
                                            n  NO1'with Oj,  ppm
                                               • NO concentration before
                                                 titration with O,, ppm
                                                NO concentration remain-
                                                ing after titration with
                                                O3, ppm
                  [NO]R  = &% scale, 0.000  ppm  (Eq.  1,  Fig 6.2)
                  [NO]A  SB &. OOP ppm
Audit point:    [NOJR  = <\I..Q % scale,  0.
                  [NO]A  = p. i\s0  ppm  (Eq 3)
                                                        ppm  (Eq  1, Fig 6.2)
Flow
measurements
Total flow
. 
t,
min
O . 3.SO
o. nHo
e>.Zf.r>
o. ar\(a
?> . on^
o . 2,*^
^vg'
min
O- 35*0
o. a^t
V,
cm
\ooo
*
\0
FT/ . cm /min
(Equation 2)
-Hooo
^G. o
    Figure  6.2   Example  of an NO  audit calculation  form.

-------
                                           Section 2.3.6
                                           Date June 5, 1984
                                           Page 6
Audit point:  [NO:|R =
              [NO]A =
                             %  scale<  Q-SLO*> ppm  (Eg 1, Fig 6'. 2)
                              ppm  (Eg 3)
Flow
measurements
Total flow

t,
min
& . 3£oO
O, 3-xo
o . 3-KO
0. (*ZS
&(*
tavg'
min.
o . 3.^0
0 >(*$£'
v,
cm
\oc?o
\0
FT/ cm /min
(Eguation 2)
•Kooo
\6>. O
Audit point:
              [NO]R =
              [NO]A"=
                               scale/  o -OS" ppm  (Eq 1)
                              ppm  (Eg 3)
Flow
measurements
Total flow
(FT)
NO flow

t,
min
 . J -^<&
o . as-o
<^>. As-0
A. ^rv<=»
-^ . •er<9o
^2, H^O
tavg/
min
o.as'o
^-S^^o
v,
cm3
\ooo
\0
F_, cm /min
(Eguation 2)
•Kexoo
M,,^
NO audit calibration eguation (y = b x +a)
NO audit concentration (x) vs. analyzer response in % scale (y)
              Slope (b) =   _?^;a .
          Intercept (a) =      ^
        Correlation (r) = 	\ .
    Figure 6.3  Example of an NO audit calculation form.

-------
                                             Section  No.  2.3.6
                                             Date June  5,  1984
                                             Page 7
Audit
point
1
2
3
- i
. *
[NO]
scale
 2.75 ppm-min,          Equation 6-1
      [NO]
          RC
                             'NO
                           O  + *NO
Equation' 6-2
                                                    Equation 6-3

-------
                                            Section No.  2.3.6
                                            Date  June  5,  1984
                                            Page  8
where
          P  = dynamic   parameter    specification/    determined
               empirically/  to ensure-  complete  reaction of  the
               available 03, ppm-min,
       [NO]_,_ = NO concentration in  the  reaction  chamber/  ppm,
          jKt.
          tR - residence time  in  the  reaction  chamber,  min,
      [NO]__ - concentration of  the NO  gas  cylinder,  ppm
         S__
                                                         ,
             = NO  flow rate,- standard  cm /min,
          FQ * 03  generator  air  flow rate,  standard  cm /min,  and
         V_c « volume of the reaction  chamber,  cm .
     The  flow  conditions to be  used  are selected according  to
the following sequence:
     a.   Determine  F_,  the total  flow  rate  required  at  the
output  manifold  (F_ =  analyzer (s) demand plus  10%  to 50%  ex-
cess)".                                                          .
     b.   Determine  F   ,  the flow rate  of  NO  required to  gen-
erate the  lowest NO concentration required at  the output  mani-
fold during the  GPT  (approximately 0.15  ppm).

                      0.15 x Fm     .                     *.•    r »
                 „   _         T                      Equation 6-4
                 FNO
     c.   Measure  the  system's  reaction  chamber  volume;  this
must be in the .range of approximately  100  to  500  cm .
     d.   Compute F  .                                .
           p  _1[N°]STD XFNO.X VRC     _
            01        2.75      '      *NO.         Equation 6-5

     e.   Compute  tp,  using  Equation  6-3;  verify  that t^  <  2
                    XV                                      J^  —
min.

-------
                                             Section No.  2.3.6
                                             Date June 5,  1984
                                             Page 9
       f.    Compute  F  .
                FD    FT  *  F0  ~  FNO,       -           Equation  6-6
 where
                FD =  diluent  air  flow, standard cm3/min.
      Adjust FQ to the value  determined  above.   FQ should not be
 further adjusted during the  NO-NOx or NO2 audit procedures; only
 FNO  (or  PD-)  and the °3 generator settings  are  adjusted during
 the course of the audit.
      9.   Connect the  analyzer  sample  line  inlet to  the glass
 manifold  of  the audit  device.  The  test atmosphere  must  pass
 through all  filters, scrubbers,  conditioners, and 'other compo-
 nents  u-sed  during  normal  ambient  sampling  and  as much  of  the
 ambient air inlet system as is practical.
     10.   Allow the  analyzer  to sample zero  air until  stable
 responses are obtained.   Record the analyzer  zero values for all
 channels  audited in  the appropriate  spaces  of the audit form,
 Figure.6.1.
     11.    Generate  the  first up-scale audit  point by  adding  the
 NO  reference  gas  to the  zero  air, making sure the generated  con-
 centrations are within one of the required  concentration ranges.
     12.    After a stable trace is obtained,  record the  NO chan-
 nel  response  [NO]R in %  scale on  the NO  calculation  form,  Figure
 6.2.   Calculate the  [NO]R,  in ppm  using Equation  1  of  Figure
 6.2,  where b and a are  the  slope and intercept  of the  station
 calibration line.                        .
     13.   Attach  the  soap bubble  meter  to  the  audit  device
 outlet  line,  and determine  the  total flow rate  (F ) .   Do not
 adjust  the  audit  device settings.  Record  the  value in Figures
 6.2 and 6.3, and calculate the  total flow rate using'Equation  2.
    14.    Disconnect  the  NO  flow line at the  audit  device  re-
action  chamber.  - Connect  the  soap  bubble meter  to  the  line-
determine  the  NO  flow rate  (FNQ)  using  Equation  2;  and record
the value on the calculation form.

-------
                                           Section No. 2.3.6
                                           Date June 5, 1984
                                           Page 10

    15.   Calculate the audit value  [NO]A for first audit point
using Equation 3, and record the results.
    16.   Repeat steps 11 through .15  for the remaining up-scale
audit points.  A minimum of three upscale  audit  points is rec-
ommended .
    17.   Transfer the NO audit concentrations and the station
NO response values to the summary form, Figure 6.1.
    18.'   Calculate • the   percentage  difference  for  each  audit
point using  Equation 4 of Figure  6.2,  and record  each on Fig-
ure 6.1.
    19.   Prepare an audit calibration  curve for  the NO channel
by using least squares-.  Include the zero air point.   (The audit
concentration is the x variable; the analyzer  response in per-
cent scale is the y variable.)  The NO calibration  curve will be
used  to  determine  the actual  audit  concentrations  during the
generation of the NO_ atmospheres.
Procedure for NO,, Audit Using Gas Phase Titration  (GPT) -
     1.   Generate  an  NO concentration which  is  approximately
0.08  to  0.12 ppm  higher than  the  NO2  audit  concentration re-
quired.  Allow the analyzer to  sample this concentration until a
stable  response  is  obtained.   Determine the  NO  concentration
[NO]—,.,,  from the  NO audit  calibration  relationship.  Record
    UKlo
this  value  for  the  first audit point  in  column 1  of the N02
calculation form, Figure  6.4.
     2.   Add O3 to reduce  the NO  concentration  by an amount
equivalent to the  NO_ audit  concentration  required.  After the
analyzer  response  stabilizes,  determine  the  NO  concentration
remaining  [NO]..-.,  'from  the  NO  audit  calibration  relationship.
              K-CiM
Record this  value  for  the first audit  point  in column 2 of the
calculation form.
     3.   Calculate  the  N0_  audit  concentration  [NO_]A using
Equation 5  of Figure 6.2 and  record  in column  3  of  the cal-
culation form, Figure 6.4.

-------
                                             Section No.  2.3.6
                                             Date June 5,  1984
                                             Page 11

       4.   Record the  station N02  response  [NO2]R in column  4.
       5.   Repeat  steps 1  through 4 for  the remaining  up-scale
 audit points.
       6.   Transfer  the N02 audit concentrations and  the  station
 N02 responses to Figure 6.1.
      7.   Determine  the percentage  difference  for each point,
 using Equation 4 of Figure 6.2.                      '
 Procedure for NO  Audit -
                 Jt""™*^^*^^^^™^^~"
      The N0x channel  may  be  audited by-the  same method used to
 audit the  NO channel and by  determining  converter efficiency
 using data from the N0x and NO- audit.
      Use Figure  6.1  to report  the audit  value,  the  analyzer
 response,  and the percentage difference.  Record the information
 on Figures  6.1,  6.2,  6.3, and  6.4.   Mark  the  recorder strip
 charts with  the  following informations •
      1.    Person  conducting the audit.
      2.    Time and  date.of audit.
      3.    Concentrations of audit points.
      4.    Identification of each audit point.
 6-1'4   Interpretation of Audit Results  -  Results  of the  audit
 will be  used  to.estimate the accuracy of the  ambient air quality
 data.  Calculation.of  accuracy  is  described  in  Subsection 2.0.8.

 6.2  Data Reduction Audit

     Data  reduction  involves  reading  a   strip  chart   record,
 calculating  an  average, and  transcribing  or recording  the  re-
 sults on the  SAROAD form.   This independent check  of  the  entire
data reduction should  be. performed by an  individual other  than
the one  who  originally reduced  the  data.    Initially the  data
processing check should  be  performed  1 day  out  of every 2 weeks
of data.  For two  1-hour periods  within each day audited, make
independent  readings   of  the  strip  chart  record  and continue

-------
                                           Section No. 2.3.6
                                           Date June 5, 1984
                                           Page 12

through the actual transcription of the data on the SAROAD form.
The 2 hours selected during each day audited should be those for
which either the trace is most dynamic, in terms of spikes or the
average concentration is high.
     The data processing check  is  made by calculating the accu-
racy ,

               A =- [N02]R - [N02]CHE(,K
where          A » the difference in measured and check values,
           [NO.,].., = the recorded analyzer response, ppm, and
             ^ R
                 - the data processing NO- concentration, ppm.
                                   >      £.

If A exceeds  ±0.02  ppm,  check all of  the  remaining  data in the
2-week period.
6.3-  Systems Audit
     A systems audit  is  an  on-site inspection and review of the
quality  assurance activities  used  for  the total  measurement
system  (sample collection,  sample  analysis,,  data  processing,
etc.);  it is a qualitative appraisal of system-quality.
     Conduct  the  systems audit  at the  startup of a  new moni-
toring system  and periodically  (as  appropriate)  as  significant
changes in system operations occur.
     A checklist  for  a systems audit  is  Figure 6.5.   Questions
in this  checklist should be  reviewed for applicability  to the
particular local,  State, or Federal agency.

-------
YES  NO

          1
         7.


         8.


         9.
        12.

        13.
                                           Section No. 2.3.6
                                           Date June 5, 1984
                                           Page 13
      Zero, span and precision checks performed  weekly  or
      1 biweekly                                       •*

  2.  Voltage and temperature variations monitored

  3.  Flow meters routinely calibrated to ±2 % accuracy
       aoains* a ,.oH»*i« ^^-^ such a£J fi .goap bubbfe
  4.  Flow rates monitored routinely

  5.  Excessive noise minimized

  6.  Data processing checks performed

      Multipoint calibrations performed routinely and'a
       record of these calibrations maintained

      Quality control charts maintained for zero and
       span checks

      Maintenance routinely  performed on pertinent
       components as  per manufacturer's manual

10.   Calibration gases traceable'to  NBS-SRM's

11.   Sample  introduction system check made weekly

      Particulate filter .(if used)  changed  frequently

      Recording system checked and  serviced before
      each sampling period

14.  Recorded data checked for. signs of system
      malfunctions


15"  DIoanqUal:Lty re?o;fds maintained—completeness,
      accuracy, precision,  and representativeness

     Figure 6.5  Checklist for use by auditor  (measurement
                 of continuous NO2 in ambient air).

-------
Audit
Multipoint
 calibration
 audit
Data audit
                                                  Section  2.3.6
                                                  Date June 5, 1984
                                                  Page 14
                 Table 6.1  ACTIVITY MATRIX FOR  AUDIT  PROCEDURE
Acceptance limits
The difference in con-
centrations between
the measured values
and the audit values
is used as a measure
of accuracy, Sec
2.0.8
Adhere to stepwise
procedure for data
reduction, Sec 2.3.4;
no difference exceed-
ing +0.02 ppm
Frequency of method
   of measurement
At least once per
quarter; Subsec
6". 1.3 for procedure
Perform independent
data processing check
on a sample of the
recorded data, e.g.,
check 1 day out
of every 2 weeks of
data, 2 hours for
each day
Action if
requirements
are not met
Recalibrate
the analyzer
Check all
remaining data
if one or more
data reduction
checks exceed
+0.02 ppm
Systems audit
Method" as described in
this section of the
Handbook
At the startup
of a new monitoring
system, and periodi-
cally as appropriate;
observation and
checklist (Fig 6.5)
Initiate
improved
methods and/
or training
programs

-------
                                             Section No. 2.3.7
                                             Revision No. 1
                                             Date July 1, 1979
                                             Page 1 of 1
7.0   ASSESSMENT OF  MONITORING DATA  FOR. PRECISION  AND ACCURACY
     For  continuous  analyzers, perform  a check  every  two weeks
to  assess the'  precision  of the data.   Use these data to esti-
mate single instrument precision as described in Section 2.0.8 of
this  volume of the  Handbook.  The  precision  check procedures
described  in  Section  2.3.3  are consistent 'with those given in
References 3.and 7.
     Estimates  of  single instrument  accuracy  for   ambient  air
quality  measurements  from  continuous • methods,  are  calculated
according to the procedure  in  Section 2.0.8.  The .performance of
the audit is described in Section 2.3.6.

-------

-------
                                             Section No. 2.3.8
                                             Revision No. 0
                                             Date July 1, 1979
                                             Page 1 of 1
8.0  RECOMMENDED STANDARDS FOR ESTABLISHING TRACEABILITY
     To  assure  data of  desired quality,  two  considerations are
essential:   (1)  the measurement  process must be  in statistical
control  at the  time  of  the  measurement and  (2)  the systematic
errors,  when  combined  with the random  variation in the measure-
ment process, must result in a suitably small uncertainty. '
     Evidence of goo~d  quality data includes documentation of the
quality control checks and the independent audits of the measure-
ment process by  recording data on specific forms or on a quality
control  chart  and  by using materials,  instruments,  and measure-
ment procedures  that  can be  traced to  appropriate  'standards  of
reference.   To  establish traceability,  data must be  obtained
routinely  by  repeat measurements  of standard reference  samples
(primary,  secondary,-and/or  working standards),   and a condition
of  process  control must be  established.   More  specifically,
working .calibration standards should be traceable to standards  of
higher accuracy,  such as  those listed below.
               NBS-SRM's AVAILABLE FOR TRACEABILITY OF
                 CALIBRATION AND AUDIT GAS STANDARDS
                        Cylinder Gases
NBS-SRM4
1683a
'1684a
1685a '
Type
Nitric oxide in N_
Nitric oxide in N9
£•
Nitric oxide in N2
Size,
S. at STP
870
870
870
Nominal .
concentration
50 ppm
100 ppm '
250 ppm
                        Permeation Tubes
NBS-SRM4
1629
Type
Nitrogen dioxide
Permeation
rate,
(jg/min
at 25°C
. 1.0
Concentration, ppm
at flow rates 'of
1 2/min
0.5
5 £/min
0.1

-------

-------
                                             Section No.  2.3.10
                                             Revision No.  0
                                             Date July 1,  1979
                                             Page 1 of 1
10.0  REFERENCES
     1.   Environmental  Protection Agency,  Title  40,  Code  of
          Federal Regulations,  Part 50  - Measurement  Principle
          and  Calibration  Procedure  for  the   Measurement  of
          Nitrogen Dioxide  in the  Atmosphere  (Gas Phase  Chemi-
          luminescence),    Federal   Register,   41.   (232):    pp
          52688-52692,  December 1976.

     2.   Ellis,  Elizabeth C.  Technical  Assistance Document for
          the Chemiluminescence Measurement of Nitrogen Dioxide,
          U.S.  Environmental Protection Agency, Research Triangle
          Park,  NC.   October 1976.   91  p.

     3.   Appendix A - Quality Assurance Requirements  for State
          and  Local  -Air .Monitoring  Stations  (SLAMS),  Federal
          Register,  Vol.  44,  .No.  92,  pp  27574-27582  May  1979.

     4.   Catalog  of  NBS  Standard Reference   Materials.    NBS
          Special Publication 260,  1975-76 Edition.  U.S.  Depart-
          ment .of  Commerce,  NBS.   Washington,  D.C.   June  1975.

     5.   Quality Assurance Handbook for Air Pollution  Measure-
          ment Systems  -  Volume I,  Principles.  EPA-600/9-76-005.
          March 1976.   .                 .  ,

     6.   Users Manual:   SAROAD  (Storage and Retrieval  of Aero-
          metric Data).  PB 201-408, APTD-0663.   July  1971.

     7.   Appendix B Quality  Assurance Requirements for  Preven-
        •  tion of Significant Deterioration (PSD)  Air  Monitoring,
          Federal Register. Vol. 44,  No. 92, pp  27582-27584 May
          1979.

-------

-------

-------
.i: i'lLiii:1!;:!!1'1!:!,;;!,1	.ilillliEiqiliiiiH

-------
                 Jan. 1983
x°/EPA
                        United States
                        Environmental Protection
                        Agency
                            Section 2.6.0

                            Environmental Monitoring Systems
                            Laboratory
                            Research Triangle Park NC 27711
                        Research and Development
                            EPA-600/4-77-027a
Test Method
Section  2.6
Reference Method  for  the
Determination of  Carbon
Monoxide  in  the Atmosphere
(Nondispersive  Infrared
Photometry)
                                              Outline.
                                    Section
                       Summary
                       Method Highlights
                       Method Description
                         1. Procurement of Equipment and
                           Supplies
                         2. Calibration of Equipment
                         3. Operation and Procedure
                         4. Data Reduction, Validation, and
                           •Reporting
                       •  5. Maintenance
                         6. Auditing Procedure
                         7. Assessment of Monitoring Data
                           for Precision and Accuracy
                         8. Recommended Standards for
                           Establishing Traceability
                         9. Reference Method
                        10. References
                        11. Data Forms
                       Summary
                        Measurements of carbon monoxide
                       (CO) in ambient air are based on the
                       absorption of infrared radiation by CO
                       in a nondispersive photometer.
                       Infrared energy from a source is
                       passed through a cell containing the
                       gas sample to be analyzed, and the
                       quantitative absorption of energy by
                       CO in the sample cell is measured by
                       a suitable detector. The photometer is
                                            Number of
                               Documentation   pages
                                  '  2.6          !
                                  '  2.6          T
                                    2.6.1
                                    2.6.2
                                    2.6.3

                                   '2.6.4
                                    2.6.5
                                    2,6.6

                                    2.6.7

                                    2.6.8
                                   .2.6.9
                                    2.6.10
                                    2.6.11
 4
 6
 6

 3
 2
 4
 1
 3
 1
12
                          sensitized to CO by employing CO gas
                          in either the detector or in a filter cell
                          in the optical path, thereby limiting
                          the measured absorption to one or
                          more of the characteristic
                          wavelengths at which CO strongly
                          absorbs. Optical filters or other means
                          may also be used to limit sensitivity of
                          the photometer to a narrow band of
                          interest. Various schemes may be
                          used to provide a suitable zero

-------
                           Section 2.6.0
                                                                             Jan. 1983
reference for the photometer. The
measured absorption is converted to
an electrical output signal, which is
related to the concentration of CO in
the measurement cell.
  An analyzer based on this principle
will be considered a Reference
Method only  if it has been designated
as a Reference Method in accordance
with 40 CFR  53.
  A current list of all designated
Reference and Equivalent Methods is
maintained by EPA and updated
whenever a new method is
designated. This list may be obtained
from any EPA Regional Office or-from
the Environmental Monitoring
Systems Laboratory. Department E,
MD-77, Research Triangle Park,
North Carolina 27711. Moreover, any
analyzer offered for sale as a
Reference or Equivalent Method after
April 16, 1976. must bear a label or
sticker indicating that the EPA has so
designated it. Further  discussion of
the concepts  of Reference and
Equivalent Methods appears in
Section 2.0.4 of this Handbook.
  Quality assurance procedures for  .
measuring CO with a  nondispersive,
infrared radiation,  automated
sampler are not instrument specific;
therefore, the following quality
assurance functions are applicable to
all CO analyzers"designated as EPA
Reference Methods.

Method Highlights
  This section presents procedures for
the Carbon Monoxide  (CO) Reference
Method {Nondispersive Infrared
Photometry),  which are intended to
serve as guidelines for the
development  of agency quality
assurance programs. Because
recordkeeping is critical in quality
assurance activities, example data
forms are included to aid in data
documentation. The blank data forms
(Section 2.6.11) may be used as they
are, or they may serve as a basis for
the preparation of forms more
appropriate to the individual agency;
the partially filled-in forms are
interspersed throughout the method
description to illustrate their uses.
Activity matrices at the end of
pertinent sections provide quick
reviews of the method description.
The CO method is  summarized briefly
in the remainder of this section.

1.  Procurement of Equipment and
Supplies
  Section 2.6,1 gives the
specifications, criteria, and design
features of the equipment and the
supplies needed for the operation of
and quality assurance checks on a
 continuous CO analyzer. Selection of
 the correct equipment and supplies is
 a prerequisite of a quality assurance
 program. This section provides a guide
 for the procurement and the initial
 checks of equipment and supplies.

 2.  Calibration of Equipment
  Section 2.6.2 provides procedures
 and forms to be used in performing a
 multipoint calibration, and in
 evaluating the calibration data.
 Subsection 2.1 deals primarily with
 minimum acceptable requirements for
 standards to be applied to the
 generation of CO concentrations.
 Subsection 2.2 provides step-by-step
 recommended calibration procedures
 for a nondispersive infrared (NDIR) CO
 analyzer, along with example
 calculations. The  data form (Figures
 2.1 and 2.2) is to be used in the
 documentation of calibration data.
 Dynamic instrument calibration is
 essential for quality control.

 3.  Operation and Procedure
  Section 2.6.3 outlines the protocol
 to be followed by the operator during
 each site visit. To provide
 documentation and accountability of
 activities,  the operator should compile
 and fill out a checklist, similar to the
 example in Figure 3.1 of Section
 2.6.3, as each activity is completed.
 Checks should include visual
 inspection of the shelter, the sample
 introduction system, the analyzer, and
 the recorder.  Level 1 zero and span
 checks must be carried out at least
 once every 2 weeks; Level 2 checks
 should be conducted between the
 Level 1 checks at a frequency
 established by the user.  Span
 concentrations for both levels should
 be between 70 and 90 percent of the
 measurement range. A one-point
 precision check should be made every
 2 weeks at a CO concentration
 between 8 and 10 ppm.  Data forms
 similar to  Figures 3.2 and 3.3 of
 Section 2.6.3 should be  used to
 document the analyzer performance
 checks. Routinely scheduled checks to
verify the operational status of the
 monitoring system are essential in  a
 quality assurance program.

 4.  Data Reduction, Validation,  and
Reporting
  Section  2.6.4 describes procedures
to be used for editing strip charts and
for data validation and reduction. Data
collected on strip charts  serve no
 useful function until they are
converted  into meaningful units
(fjg/m3. ppm) applying to a specific
time period (e.g., hourly) through the
use of the calibration relationship.
These data must be transcribed  into
 an appropriate format such as that of
 the SAROAD hourly data form.

 5. Maintenance
   Section 2.6.5 addresses the
 recordkeeping and the scheduled
 activities pertinent to preventive and
 corrective maintenance. A sample
 maintenance log is shown in Figure
 5.1. Preventive and corrective
 maintenance are necessary to
 minimize loss of air quality data due
 to analyzer malfunctions and out-of-
 control conditions.

 6. Assessment of Data for Accuracy
 and Precision
   Section 2.6.6 discusses procedures
 and forms for system and
 performance audits.  Multipoint
 performance audits to be used to
 assess the accuracy of the data
 collection are discussed in Subsection
 6.1.1; audit procedures are given in
 Subsection 6.1.2; a data reduction
 audit is discussed in Subsection 6.1.3;
 and a system audit is discussed in
 Subsection 6.2. Figure 6.1 presents
 examples of audit summary and audit
 calculation forms. Figure 6.2  is an
 example checklist to be used  by the
 auditor.
   Section 2.6.7 describes the
 techniques for assessment of data for
 accuracy and precision.

 7. Reference Information
   Section 2.6.8 discusses the
 traceability of standards to established
•standards of higher accuracy, a
 prerequisite for  obtaining accurate
 data.
   Sections 2.6.9 and 2.6.10 contain
 the Reference Method and pertinent
 references.

-------
                               Jan. 1983
                                                                              Section 2.6.1
                           1.0   Procurement of Equipment and Supplies
     Measurement of carbon monoxide
   (CO) in ambient air requires basic
   sampling equipment and peripheral
   supplies; these include, but are not
   limited to. the following:
     1.  Reference method CO analyzer
        (NDIR) (Subsection 1.1 provides
        information on obtaining an up-
        to-date list of analyzers)
     2.  Strip chart recorder or data
        logging system
     3.  Sampling lines
     4.  Sampling manifold
     5.  Calibration equipment
     6.  NBS-SRM or  commercial CRM
        calibration standard
     7. Working calibration and audit
       gases traceable to NBS or CRM
       standard
     8. Zero-air source
     9. Spare parts and.expendable
       supplies
   10. Record forms
   11. .Independent audit system.
     The person responsible for
  purchasing materials should maintain
  a log  to record vendor names, part
  numbers,  prices, dates, and other
  pertinent information. An example log
  is  shown in  Figure 1.1. The log will
  serve as a reference for- future
  procurement needs and as a tool for
  planning budgets for future
  monitoring programs. Quality
  assurance activities for procurement
  of  equipment and supplies are
  summarized in Table 1.2 at the end of
  this section.

  1.1   The CO Analyzer (NDIR)
   As stated in the Code of  Federal
  Regulations,' each  method for
  measuring CO shall be either a
  Reference  or Equivalent Method when
  the purpose .is to determine
  compliance with the National Ambient
  Air Quality Standards (NAAQS). For
  carbon monoxide, the Reference
  Method is Nondispersive Infrared
  Photometry (NDIR).
   Although the NDIR analyzers
 currently available for measuring CO
 in ambient  air are competitively
 priced, price differences become
 apparent when options to the basic
 package are ordered. The buyer
 should consult the list of designated
 Reference and Equivalent Methods for
lapproved options. An up-to-date list of
'analyzers designated as reference  or
 Equivalent methods  for  CO is
 available by writing to:
     U.S. Environmental Protection Agency
     Environmental Monitoring Systems
     Laboratory
     Department E, MD-77
     Research Triangle Park, North
     Carolina 27711
  Available options include automatic
  zero and span systems and complete
  telemetry systems for  transmitting
  daily zero and span checks and real-
•  time data from the site to a central
  location. For certain CO analyzers, the
  automatic zero and span systems are
  required to meet the EPA Reference   '
  Method designation. Although options
  can add convenience and flexibility,
  their necessity and desirability must
  be dictated by the availability of field
  personnel, accessibility of the site,
  and limitations of the budget.
    When purchasing, the buyer should
  request that the manufacturer supply
  documented proof that the specific
  analyzer performs within
  specifications  (Table 4.1, Section
  2.0.4). The best proof is a strip chart
  recording showing the  specific
  analyzer's zero drift, span drift,
  electronic noise, risetime, falltime,
  and lagtime. Acceptance of the  •
  analyzer should be based on these
  performance tests; once -accepted, the
  Reference and Equivalent analyzers
  are warranted by the manufacturer to
  operate within the required
  performance limits for 1 year. The
 strip chart will also serve as a
 reference to determine  whether the
 performance of the analyzer has
 deteriorated at a later date. The user
 should reverify the performance
 characteristics either during the initial
 calibration or by using abbreviated
 forms of the test procedures in the
 ambient air  monitoring  Reference and
 Equivalent Methods Regulations.2

 1.2  Strip Chart  Recorder
  Recorders are commercially
 available at a wide variety of prices
 and specifications. Factors to  be
 considered in the purchase of a
 recorder are:
  1. Compatibility with the output
     signal of the analyzer
  2. A minimum chart width of 15
     cm (6 in.) for the desired
     accuracy in data reduction
  3. A minimum chart speed of at
     least 2.5 cm/h (1 in./h)
  4. Response time
  5. Precision and reliability
    6.  Flexibility of operating variables
       (speed, range)
    7.  Maintenance requirements.

  1.3   Sampling  Lines  and
  Manifold
   Wherever possible, sampling lines
  and manifolds should be constructed
  of Teflon or glass to  minimize
  degradation of the sample;  however,
  because of the relative inertness of
  CO, other types of materials
  (polypropylene, stainless steel) will
  suffice if only CO is being measured.
  Sample residence time should be
  minimized. The use of a particle filter
  on the sample inlet line of an NDIR
  CO analyzer is optional on some
 'analyzers, and left to the discretion of
 the-user or the manufacturer. Use of
 the filter should depend on  the
 analyzer's susceptibility to
 interference, malfunction, or damage
 due to particles.

  1.4  Calibration  Equipment
 and Standards
   The two acceptable methods for
 dynamic multipoint calibration of  CO
 analyzers are:3
   1. The use of individual certified
 standard cylinders of  CO for each
 concentration needed.
   2. The use of one certified standard
 cylinder of CO, diluted as necessary
 with zero-air, to obtain the various
 calibration concentrations needed.
   Both methods require the
 following:
   1. Pressure regulator(s) for CO
 cylinder(s)
   2. Flow controller
   3. Flow meter
   4. Mixing chamber  (dynamic  .
 dilution only)
   5. Output manifold           •
  •6. Zero-air source
   7. Calibration standard.
  The equipment needed for calibration
can be purchased commercially, or it
can be assembled by the user. When
a calibrator or its components are
being purchased, certain factors must
be considered:
  1.  Traceability of the certified
     calibration gases  to an NBS-
     SRM4 or an  NBS/EPA-approved
     commercially available Certified
     Reference Material  (CRM).
  2. Accuracy of the flow-measuring
    device (rotameter, mass flow
    meter, bubble meter).

-------
              Section 2.6.1
                                                         Jan. 1983
    
-------
                               Jan. 1983
                                                                              Section 2.6.1
    3.  Maximum and minimum flows of
       dilution air and calibration gases.
    4.  Ease of transporting the
       calibration equipment from site
       to site.

  1.4.1  Pressure Regulator—
    A pressure regulator will be
  required for the CO calibration
  standard cylinder.  If individual
  cylinders  are to be used for individual
  calibration points,  it is advisable to
  procure regulators for each cylinder.
  Regulators must have a nonreactive
  diaphragm and suitable delivery
  pressure.  A two-stage regulator with
  inlet  and delivery pressure gauges is
  recommended. The .supplier from
  which the CO cylinders are to be
  obtained should be consulted as to
  Ihe correct cylinder fitting size
  required for the regulator.

  1.4.2 Flow Controller —
    The flow controller can be any
  device (valve) capable of adjusting and
  regulating the flow from the
  calibration standard. If the dilution
  method is to be used for calibration, a
  second device will  be required for the
  zero-air. For dilution, the  controllers
  must be capable of regulating the flow
  to ±1 percent.

  1.4.3 Flow Meter —
   A calibrated flow meter capable of   •
.  measuring and monitoring the
  calibration standard flow  rate  will be
  required. If the dilution method is
  used, a second flow meter will be
  required for the zero-air flow.  For
  dilution, the flow meters must be
  capable of measuring the flow with an
  accuracy of ±2 percent.

  1.4.4  Mixing Chamber —
   A mixing chamber is required  only if
 the calibrator concentrations are
 generated  by dynamic  dilution of a CO
 standard. The chamber should be
 designed to provide thorough mixing
 of CO and  zero-air.

  1.4.5  Output Manifold —
   The output manifold should  be of
 sufficient diameter  to insure an
 insignificant pressure drop at the
 analyzer connection. The system
 must have a vent designed to  insure
 atmospheric pressure at the manifold
 and to prevent  ambient air from
 entering the manifold.

 1.4.6  Zero-Air Source —
   A source of dry zero-air that is
 verified to be'free of contaminants
 that could cause detectable responses
 from the CO analyzer will  be needed.
 Zero-air containing  <0.1 ppm CO may
 be purchased in high-pressure
 cylinders or generated with
 commercially available clean air
 systems. The zero-air must contain
 
-------
                        Section 2.6.1
                                                                          Jan. 1983
Table 1.2.    Activity Matrix for Procurement of Equipment and Supplies
Equipment and
supplies	
NDIR analyzer
Strip chart recorder
Sampling lines and manifold
Calibration gases
Audit gases
Zero-air
Acceptance limits
Performance according to
specifications in Table 4.1,
Sec. 2.0.4
Compatible with output
signal of analyzer: recom-
mended chart width of 15
cm (6 in.)
Constructed of Teflon or
glass

Traceable to an NBS-SRM
or a commercially available
CRM; ±2.0% of rated
concentration

Traceable to an fJBS-SRM
or commercially available
CRM: ±2.0% of rated
concentration
<0.1 ppm CO
Frequency and method
of measurement
Manufacturer strip chart
recording of analyzer's
performance
Visually observe upon
receipt
Visually observe upon
receipt

Upon receipt and monthly
thereafter for first 3 mo: if
concentration remains'
stable,  verify every 6th
month
As above
Check against analyzer
internal zero or another
source of zero-air known to
to be CO-free
Action if
requirements
are not met
Have the manufacturer
adjust and rerun the per-
formance checks
Return to supplier
Other types of materials
may be acceptable for CO
sampling •
Return to supplier
As above
As above

-------
                               Jan. 1983
                                                                               Section 2.6.2
                                     2.0   Calibration  of Equipment
    The accuracy and validity of
  measurement data recorded by air
  monitoring equipment depend on the
  quality assurance procedures used.
  The primary procedure is dynamic
  calibration, which determines the
  relationship between the observed
  and the actual values of the variable
  being measured.
    In dynamic multipoint calibration,
  gas samples of known concentrations.
  are introduced to an instrument to
  derive a calibration relationship or to
  adjust the  instrument to a
  predetermined sensitivity. The
  relationship is derived from the
  instrumental responses to the
  successive samples of known
  concentrations. A minimum  of four
  reference points and a zero point are
  recommended to derive the
  relationship. The "true" value of each
 calibration  gas must be traceable to
 an NBS-SRM or a commercially
 available CRM (Section 2.Q.7)..
   Most present-day monitoring
 systems are subject to drift and
 variability o£ internal parameters, and
 they cannot be expected to.maintain
 calibrations over long periods of time.
 Therefore, it is necessary that the
 calibration relationship be dynamically
 checked on a predetermined  schedule.
 Precision is determined by a  one-point
 check performed at least once every 2
 weeks. Network accuracy is determined
 by a three-point audit performed at
 least once each  quarter. Zero and
 span checks must be made to
 document in-control conditions; these
 checks are also used in data reduction
 and validation.
   Table 2.1  at the end of this section
 summarizes the  quality assurance
 activities for calibration procedures.

 2.1   Calibration  Gases

 2.1.1   CO Standard —
   The CO standards must be  in air
 unless the dilution method is used.
 For dilution, CO in nitrogen may be
 used if the zero-air dilution ratio is not
 less than 100:1.  All calibration gas
 mixtures must be referenced against
 an NBS-SRM or a commercially
 available CRM (Section 2.0.7). The
 steps required for comparing the
concentration of  a commercial
working calibration standard to the
concentration of  an NBS-SRM or a
CRM are described in Subsection 7.1
of Section 2.0.7.  Subsections  7.1.4
  and 7.1.5 describe the procedures for
  verification and reanalysis of cylinder
  gases. The CO gas cylinders should be
  recertified every 6 months. The use of
  aluminum cylinders will provide better
  stability of CO standards.

  2.1.2  Dilution Gases —
    Zero-air, verified to be free of
  contaminants that would cause
  detectable responses in the CO
  analyzer, may be purchased in  high-
  pressure cylinders or generated with
  commercially available clean air
  systems. Care must be exercised to
  ensure that <0.1 ppm CO is present
  in the zero air; some air-cylinders sold
  as ultrapure may actually contain 1  to •
  2  ppm CO. Any zero air source  used
  must be  verified to contain <0.1 ppm
  CO. The  use of a catalytic oxidizing
  agent such as Hopcalite on any zero-
  air source would be prudent. Any zero
  air passing through a catalytic oxidizer
  must be  free of water vapor.

 2.2  Calibration
   The procedure for dynamically
 calibrating the NDIR analyzer may be
 found in  40 CFR'50,3 and in the
 manufacturer's manual.  Essentially,
 the procedure involves challenging
 the analyzer with a minimum of four
 CO concentrations  and defining  the
 relationship between the
 concentration and the analyzer
 response. Forms  for recording
 operational and calculation data  have
 been developed to aid in the
 documentation of calibrations and
 quality assurance checks.
 Documentation of all data on the
 station, instrument, calibrator,
 reference standard, and calibration
 activity is of prime importance
 because the validity of the data
 collected  by the monitor' depends on
 its  calibration.

 2.2.1  Calibration Procedure —
  The following calibration procedure
 is based on dynamically diluting  a
 high CO concentration with zero-air.
An  alternative procedure is to use
individual cylinders containing the
desired CO concentrations, which
eliminates the necessity of dilution.
  Any dynamic dilution system used
for  calibration  must be capable of
measuring and controlling flow rates
to within ±2 percent of the required
flow. Flow meters must be calibrated
under the  conditions of use against a
  reliable standard, such as a soap
  bubble meter or a wet test meter. All
  volumetric flow rates should be
  corrected to 25°C (77°F) and 760 mm
  (29.92 in.) Hg. If both the CO and the
  zero-air flow rates are measured with
  the same device  under the same
  conditions of temperature and
  pressure, the STP correction factor in
  the calibration equations can be
  disregarded.-
    The following step-by-step
  procedure uses a data form (Figure
  2.1) to aid in the collection and
  documentation of calibration data. The
  calibration equations in Figure 2.1,
  the CO calibration and linearity check
  table,  and the calibration relationship
  plot in Figure 2.2 are given to
  facilitate the systematic recording of
  data derived during the calibration of
  the NDIR CO analyzer. The user
  should consult the manufacturer's
  manual before beginning the
  calibration because the zero and
  calibration procedures and adjustments
  differ from analyzer to analyzer.
    1. Record the official name and
 .address of the individual station. Note:
  Where appropriate, the station name,
  address, and SAROAD ID should be the
  same as that on the hourly average
  data form  (Figure 4.1 of Section 2.6.4)
  to help eliminate confusion on the
  part of persons not familiar with the
 station.
   2. Identify the analyzer being  -
 calibrated  by recording  the
 manufacturer's name, model, and
 serial number.
   3. Identify the person performing'
 the calibratioo and give the date of
 calibration.
   4. Identify the calibrator or dilution
 system used.'If the system was
 purchased, record the manufacturer's
 name, model, and serial number; if
 the system was assembled by the
 user, assign  it an identification
 number so that calibrations can be
 referenced to that particular
 apparatus.
  5. Identify, by supplier and cylinder
 number, the reference standard used;
 record the concentration of calibration
 gases determined by the user and the
 cylinder pressure; provide a record of
 NBS-SRM or CRM traceability for any
 cylinder used in a calibration; and
 include the date and the name of the
person who conducted the traceability
test. Note: Cylinders with pressures of

-------
                          Section 2.6.2
                                                                             Jan. 1983
 L Station
                                                  Calibration Summary

                                                 	 2. Analyzer
                    North
                                                                    Model
3. Calibration by

4, Calibrator mfr.

         Model

5. CO standard _
                                                            Date
                                                                  II-/0-71
                                                            S/N

   Verified against N8S-SRM
                                                            Concentration
                                 CU MKlafir CC A^-T / Cylinder pressure 5OO PS f

                                                                  9-
 6. Flow-measuring device

 7. Barometric pressure
                                                            Traceability
 8, Analyzer's sample How rate

 9, Zero knob setting
                              /
                                                            Shelter temperature
                                                            Span knob setting
                                                 Calibration Equations
Equation 2-1
Fo OfFcosF (STP correction factor]
Equation 2-2
STP Correction factor = Bf>  x    29g
                    76O    AT+273
Equation 2-3
fCGJoor * (COlrg*fw
          Fo » Fco
Equation 2-4
% scale * [COlout x 10O * 2cQ
             URL


Figure 2.1.    Example of calibration data form.
                                                            F = uncorrected flow rate for dilution air or CO standard gas. l/min
                                                            F'o = flow rate of dilution air corrected to 25°C and 760 mm Hg. l/min


                                                            Fco = flow rate of CO standard corrected to 25°C and 760 mm Hg.
                                                                l/min

                                                            BP = barometric pressure, mm Hg

                                                            AT - temperature of gas being measured, °C


                                                          [COJour = concentration at the output manifold, ppm

                                                          [CO]sTo= concentration of the .undiluted standard, ppm
                                                            Zoo = recorder response to zero air
<200 psig should not be used for
calibration because gases in cylinders
may become unstable for some
concentrations at low pressures
(Section 2.0.7).
  6, Identify the  flow-measuring
device used, and document the
traceability of its accuracy.
  7, Record the barometric pressure
and the shelter temperature before
the calibration.
  8, Record the analyzer's sample
flow rate.
  9, Record the zero and span knob
settings after the calibration so that
these settings can be used later to
determine changes in instrument
performance.
                                         Figure 2.2 contains a CO calibration
                                       and linearity check table and a graph to
                                       facilitate the plotting of the calibration
                                       data. The equations at the bottom of
                                       Figure 2.1 are to be used to obtain the
                                       entries in the table in Figure 2.2. The
                                       detailed steps of the calibration
                                       procedure are given below. Analyzer
                                       responses in these steps refer to
                                       recorder responses. The
                                       manufacturer's instrument manual
                                       should be consulted for analyzer-
                                       specific calibration procedures.
                                         1.  Select the operating range of the
                                       analyzer to be calibrated by referring
                                       to the manufacturer's manual for the
                                       ranges over which the analyzer is
                                       considered to be a reference method.
  2. Connect the recorder output
cable(s) of the analyzer to the input
terminals of the strip chart
recorder(s). All  adjustments to the
analyzer should be based on the
appropriate strip chart readings. Note:
When data acquisition systems are
used to store and/or transmit data to
a base station,  some provision must
be made to verify the accuracy of the
transmitted data. In these cases, a
voltmeter .or recorder can be used to
take readings and to make
adjustments onsite. A comparison
check must then be made between
signal outputs from the analyzer and
data received at the  base locations.
  3. Adjust the zero-air flow to the

-------
                                   Jan. 1983
                                                                                           Section 2.6.2
Calibration
points
Zero
80% URL
1
2
3

1
Fco,
1/min
(Eg 2-1 &2-2I*

0.^00
0.500
0.500
0. 500

2
Fa
1/min
(Eq2-1 &2-2I

2. (eZ-5
3.^7
5.150
12.. 000

3
[CO]OUT.
ppm
(Eg 2-31
o
HO
30
ZO
IO

4
% scale
(Ed 2-4)
5*7*
•riS^o
t>5%
45ti
?J5%

  'Equations 2-1 through 2-4 are given in Figure 2.1 and in the text.
                              100
Slope Ibl of calibration relationship
2..0
 Intercept fat of calibration relationship
Figure 2.2.    Example of calibration data form. (Linearity Check and .Calibration Relationship)

-------
                          Section 2.6.2
                                                                            Jan. 1983
 arwfyzer; the flow must exceed the
 total demand of the analyzer
 connected to the output manifold to
 ensure that no ambient air is pulled
 into the manifold vent.
   4. Allow the analyzer to sample the
 zero air until a stable response is
 obtained; adjust the analyzer zero
 control to within ±0.5 ppm of zero
 base line; and record the stable zero-
 air response (% scale) under column 4
 of the calibration table in Figure  2.2.
 /Vote: Offsetting the analyzer zero
 adjustment to +5% of scale is
 recommended to facilitate observing
 n«ga«ive zero drift. On  most analyzers,
 rtiis should be done by offsetting the  .
 recorder zero,
   5. Determine the 80 percent upper
 range limit (URL) of the analyzer.
 Example; For an analyzer with an
 operating range of 0 to SO ppm, the
 SO percent URL value would be 0.80
 x SO. or 40 ppm.
   6, Adjust the CO flow from the
 standard CO cylinder to generate a
 CO concentration of approximately 80
 percent of the URL. Measure the CO
 How, correct it to STP, and record
 unefer column 1 (Fco), on  the 80
 percent URL line.

     Fco s F x (STP correction factor)
                         Equation 2-1

 STP correction factor = BP  x   298
                       760  AT + 273
                       ' Equation 2-2
 where
   Fco = flow rate of  CO standard
       corrected to STP, l/min
     F = uncorrected flow rate, l/min
   8P=barometric pressure, mm Hg
   AT » temperature of gas being
       measured, °C
    e.- If wet test meter or bubble
 meter is used for flow measurement,
 fhe vapor pressure of water at the
 temperature of the meter must be
 subtracted from the barometric
 pressure.
   Measure the dilution air flow,
 correct it to STP, and record under
 column 2 (Fo).

    Fo - F x (STP correction factor)
   7 Calculate the CO concentration
 [.Colour using  Equation 2-3.

                JCOlsTpxFro
        (COlour =   Fo +Fco
                        Equation 2-3

Record this value on  the 80% URL
line under column  3.
  8, Calculate the  required recorder
response for span adjust (80% URL)
using Equation 2-4.
     % scale = ([COlouT x 100 U Zco
              \ URL       /
                         Equation 2-4
Allow the analyzer to sample until the
response is stable; adjust the analyzer
span until the required response is
obtained, and record the CO recorder
response on 80 percent URL line
under column 4. Note: If substantial
adjustments of the span  control are
necessary, recheck the zero and span
adjustments by repeating steps 4 and
8.
  9. After the zero and 80 percent
URL points have been set, without
further adjusting the instrument,
generate three approximately evenly
spaced points between zero and 80
percent URL by increasing the dilution
flow (Fo,) or by decreasing the CO
flow (Fco). For each concentration
generated,  calculate the CO
concentrations (using Equation 2-3)
and record the results for each point
under the appropriate column  in the
table in Figure 2.2.
  10.  On the blank graph of Figure
2.2, plot the analyzer responses
expressed in percent scale at the
recorder (y-axis) versus the
corresponding calculated
concentrations (x-axis) to obtain the
calibration  relationship. Determine the
straight line of best fit by the method
of least squares (Volume I, Appendix J
of this Handbook) using a
programmed calculator or the
calculation  data form (Figure 2.3).
Note: Because manual calculations
(using the data form) require
considerably more time than the use
of a programmed calculator, it is
•suggested that the latter  be used
when possible.
  11; After determining the slope (b)
and the intercept (a) where the line
crosses the y-axis, draw the fitted line
as follows:  On the y-axis, plot the y
intercept, a; use the equation Y = a +
bx to calculate the predicted Y value
using the 80 percent URL
concentration for the x value as the
second point on the graph. Draw a
straight line through these two points
to give a best-fit line, as shown in
Figure 2.4.
  12. After drawing the best-fit line,
determine if the analyzer response is
linear, that is, no calibration point
varies from the best-fit line by more
than 2 percent of full scale. Make a
simple test for linearity by plotting a
point 2 percent of scale above and 2
percent of scale below the point
where the best-fit line crosses the 40-
ppm level and the  10-ppm level, and
then draw a straight line  through the
+2 percent  points and one through the
-2 percent points (Figure 2.4). The two
lines (above and below the best-fit
line) define the limits between which
the calibration points can fall for the
calibration curve to be considered
linear. Points outside these limits
should be repeated to check for
calibration point errors;  if the repeated
points still fall outside the limits,
consult the manufacturer's manual
to determine and correct the problem.

2.2.2  Calibration Frequency —
  To ensure accurate measurements
of the CO concentrations, calibrate the
analyzer  at the time of installation,
and then recalibrate it as specified in
the instrument manual or:
  1. No later than 3 months after the
    most recent calibration or
    performance audit. If
    performance audit results are
    satisfactory,  recalibration must
    be performed immediately.
  2. After an interruption of more
   . than a few days in analyzer
    operation, after any repairs that
    might affect its calibration, after
    physical relocation of the
    analyzer,  or after any other
    indication (including excessive
    zero or span drift) of possible
    significant inaccuracy of the
    analyzer.  Following any of these
    activities, a Level 1 zero and
    span check should  be made to
    determine if recalibration  is
    necessary. If the zero and span
    drifts do not exceed the limits
    (Table 9.1, Section 2.0.9,
    Subsection 9.1.3), a calibration
    need not  be  performed. If either
    the  zero or the span drift exceeds
    its limit, investigate the cause of
    the  drift, take corrective action,
    and calibrate the analyzer.

-------
                                  Jan. 1983
                                                                                         Section 2.6.2
Calibration
point
Zero
80% URL
1
2
3
Concentration.
ppm
X
o
VO
30
2-0
/£
X2
0
fcOQ
900
WO
/£>O
Recorder
reading,
% scale
y
5
85
tf
45
Z.5
y2
25
72^5
40tfS
£0-25
£Z5
xy
o
31 oo
JJ5O
900
250
n - number of calibration points.
The eouation of the line fitted to the data is written as:
Y = Y * b(x-x) = (y-tixj + bx = a + bx
where Y = predicted mean response for corresponding x
b - slope of the fitted line
a = intercept where the line crosses the y-axis
a =y--
y = I///? = .
                                                                                                                              and
Figure 2.3.    Calculation form for the method of least squares.

-------
                          Section 2.6.2
                                                                             Jan. 1983
     100
      80
                                       Limits for instrument
                                       linearity check. ±2%
                                  20           30

                                    fCOJour. ppm

 Figure 2.4.   Example of a CO calibration relationship.

                "V


 Table 2.1.    Activity Matrix tor Calibration Procedures
Calibration
activities
                                Acceptance limits
                             Frequency and method
                             of measurement
                             Action if
                             requirements
                             are not met
Dilution gas



Span gases




Multipoint^ calibration
Zero-air free of contami-
nants (Sec. 2.0.7, Subsec.
7.1)

Cylinder gases certified to
NBS or CRM standard;
cylinder pressure >200
psig   .

According to calibration
procedure fSubsec.  2.2)
and data recorded (Figs 2.1
and 2.2)
Compare the new zero-air
against source known to be
free of contaminants

Assay against an NBS-
SRM or CRM semiannually
(Sec. 2.0.7)
                                                             Calibrate at least once.
                                                             quarterly;.anytime a Level
                                                             1 span check indicates dis-
                                                             crepancy; after mainte-
                                                             nance that may affect the
                                                             calibration (Subsec. 2.2)
Return to supplier, or take
corrective action with
generation system as
appropriate
Working gas standard un-
stable, and/or measurment
method out of control;
take corrective action,
e.g., obtain new span gases
Repeat the calibration

-------
                              Jan. 1983
                                                                              Section 2.6.3
                                     3.0   Operation and Procedure
    A routinely scheduled series of
  checks to verify the operational status
  of the monitoring system is an
  essential part of the quality
  assurance program. The operator
  should visit the site at least once each
  week, and he/she must make a Level
  1 zero and span check on the analyzer
  at least once every 2 weeks. The user
  may decide on the frequency of any
  Level 2 zero and span checks. In
  addition, an independent precision
  check between 8.0 and 10.0 ppm
  must be carried out at  least once
  every 2 weeks.
    Table 3.1  at the end  of this section
  summarizes the quality assurance
  activities for routine operations
  discussed in the following subsections.
  To provide documentation and
  accountability of activities, the field
  operator should compile and fill out a
  checklist as each activity is completed;
  Figure 3.1 is an example checklist.
    In Subsections  3.1 and 3.2.
  reference is made to the sampling
  shelter and sample inlet system. The
  design and-construction of these
 .components of the sampling system
  are not within the scope of this
  document, but an in-depth study of
  these is provided  in Reference 5.

  3.1   Shelter
   The shelter's role in quality
  assurance is to provide a
  temperature-controlled  environment
  in which the sampling equipment can
 operate at optimum  levels of
 performance. The mean shelter
 temperature should be between 22°
 and 28°C (72° and 82°F). A
 thermograph should be installed at the
 shelter so that daily  temperature
 fluctuations can be continuously
 recorded. Fluctuations greater than
 ~2°C (4°F) may cause the electronic
 components of the analyzer to drift
 and introduce error into the data;
 thus, fluctuations outside of the
 specifications should be identified,
 and the data for the affected time
 period should be flagged to indicate-
 possible discrepancies. Excess
 vibrations will cause analyzer
 fluctuations and should be minimized
 where possible.

 3.2  Sample Introduction
 System
  The sample introduction system
consists of an intake  port, paniculate
   and moisture traps, a sampling
   manifold, a blower, and a sampling
   line to the analyzer. As part of the
   quality assurance program, the field
   operator should inspect each of these
   components for breakage, leaks, and
   buildup of paniculate matter or other
   foreign materials; check for  moisture
   depostion in the sampling line or
   manifold; and check the connection
   between the sampling line and the
   manifold. Any component that is not
   within tolerance should be cleaned or
-   replaced immediately (Section 2.0.2).

   3.3   Recorder
     During each visit to the monitoring
   site, the field operator should check
   the recorder against the following list:
   •  1. Legibility of the ink trace
     2. Ink supply in the reservoir
     3. Chart paper  supply
     4. Chart speed control  setting
     5. Signal input range setting
     6. Time synchronization. Mark chart
   with correct time and  date.
   Any operational parameter that is not
   within tolerance must be corrected
   immediately.                  "

   3.4   Analyzer
    The user should read thoroughly the
  specific instructions, in the
  manufacturer's manual before
  attempting to operate the analyzer.
  As part of the quality assurance
  program,  each site visit should include
  a visual inspection of the  external
  parameters of the analyzer, the zero
  and span checks, and a biweekly
  precision point check.

  3.4.1   Visual Inspect/on —
    The field operator should inspect
  the external operating parameters of
  the analyzer; these will vary from
  instrument to instrument,  but they
  generally will include the following:
    1. Correct setting of  flow meter and
      regulators
   2. Cycling of temperature control
      indicators
   3. Verification that the analyzer is
      in the sampling mode rather  than
      in the zero or calibration modes
   4. Zero and span potentiometers set
      and locked at  proper values.
 3.4.2  Zero and Span Checks —
   Interim zero and span checks on the
 responses of the instrument to known
 concentrations must be used to
 document within-control conditions. If
  a response is outside of the
  prescribed limits, the analyzer is
  considered out of control, and the
  cause must be determined and
  corrected.  A quality control chart can
  be used to check the analyzer visually
  for within-control conditions.
    Level 1  and Level  2 span checks
  must be conducted in accordance with
  the specific guidance given in
  Subsection 9.1  of'Section 2.0.9. If
  permitted by the associated operation
  or instruction manual, a CO analyzer
  may temporarily operate during the
  zero and span checks at reduced vent
  or purge flows, or the test atmosphere
  may enter  the analyzer at a point
  other than  the normal sample inlet if
  the analyzer's response is  not likely
  to be altered by these deviations from
  the normal operational mode. Because
  variability information may  not be
  uncovered  by checking only part of
  the analyzer's sample-handling
  system, however, it is' recommended
  that these operational deviations be
  used only for Level 2 checks.
   Level 1 zero and span checks must
  be conducted every 2. weeks; Level 2  .
 checks should be conducted in
 between the Level  1  checks at a
 frequency decided on by the user.
 Span concentrations  for both levels
 should be between 70 and 90 percent
 of the measurement range.  The data
 should be recorded on a zero and
 span checks form such as that shown"
 in Figure 3.2.
   Level .1 zero and  span data are used
 for the followmg:

   1. To adjust the analyzer for zero
     and span drifts
   2. To decide when to calibrate the
     analyzer
   3. To decide when to invalidate
     monitoring data.
Items 1 and 2  are detailed in
Subsection 9.1.3 of Section 2.0.9;
Item 3,  in Subsection  9.1.4 of the
same section.
  When the response from a span
check is outside of the control  limits,
the cause for the extreme drift must
be determined and corrective action
taken. Some of the causes for drift are
listed below:

   1. Lack of preventive maintenance
  2. Fluctuations in electrical power
supply
  3.  Fluctuations in flow
  4.  Change in zero-air source

-------
                           Section 2.6.3
                                                                                  Jan. 1983
Site identification

Site location  J.

Sue address  -3***^.
                         nO(
                                                               Date
                                                               Technician
          t. Inspect thermograph for temperature variations greater than ±2°C (4°F); identify time frame of any temperature level out of
            tolerance
                                                                     /£.(/€,!.<>
            Comments: 1^VY)f.  (d/fa

   \S    2, Inspect sample introduction system for moisture, paniculate buildup, foreign material, breakage..leaks
           •• Comments:
         3. Is sample line connected to manifold?

            Comments:
          4. Inspect data recording system
              Legibility of trace
              Ink supply
              Paper supply
              Chart speed
              Signal range
              Time synchronization  ,
                                                                                   Corrective
                                                                                  action taken
            Comments:
^
^L

  /
         5. Inspect analyzer operational parameters

            • Sample flow rate
            • Oven temp light flashing
            • Analyzer in sampling mode
            • Zero and span potentiometers locked at correct settings
            Comments,'
                                                                           OK
                                                                                   Corrective
                                                                                  action taken
         6 Zero the analyzer

         7 Is unadjusted zero within tolerance?

            Comments:     ^&TO   O
                                                        *4. ^
 
-------
                               Jan. 1983
                                                                                  Section 2,6.3
  Site identification   ^~ LJ IJ





  Location





  Address
                                          Pollutant     C O




                                          Analyzer
AJorrti
  Adjusted zero
                                                              Serial number
                                          Adjusted span
/ /
                                                                              / J
Date
// ~/5~7^


Operator
^




Unadjusted
zero.
% chart
5-6


Span
concentration,
ppm
35>m




.
\
-~










i

' i

Unadjusted
analyzer
response. Difference.
% chart
757



ppm ppm
35 O
\
i
i
i i
i t
i
i

1 • . •
t
j
j ' •
!


I

Figure 3.2.    Example of a Level / zero and span check data form.

-------
                         Section 2.6.3
                                                                           Jan. 1983
   5. Change in span gas
 concentration
   6. Degradation of detector
   7. Electronic and physical
 components not within
 manufacturer's specifications.
 Corrective actions for the above can
 be found in the manufacturer's
 instruction/operations manual.

 3.4.3  Precision Check —
   A periodic precision check is used
 to assess the data. A one-point check
 on each analyzer must be carried out
 at least once every 2 weeks at a CO
 concentration between 8 and 10 ppm.
 The analyzer must  be operated in its
 normal sampling mode, and the
 precision test gas must pass through
 all filters, scrubbers, conditioners, and
 other components used during normal
 ambient sampling.  If permitted by the
 associated operation or instruction
 manual, a CO analyzer may
 temporarily operate during the
 precision check at reduced vent or
 purge flows, or the test atmosphere
 may enter the analyzer at a point
 other than the normal sample inlet  if
 the analyzer's response is not likely to
 be altered by these deviations from
 the normal operational mode. The
 standards from which the precision
 check test concentrations are obtained
 must be traceable to an NBS-SRM or
 a commercially available CRM; the
 standards used for  calibration may be
 used for the  precision check.
   The precision check procedure is  as
 follows:
   1. Connect the analyzer's sample
     inlet line to a precision gas
     source that has a concentration
     between 8 and 10 ppm CO and
     that is traceable to an NBS-SRM
     or a CRM.  If a precision check  is
     made in conjunction with a
     zero/span  check, it  must be
     made prior to any zero and span
     adjustments.
   2. Allow the analyzer to sample the
     precision gas for at  least 5 min
     or until  a stable recorder trace is
     obtained.
   3. Record this value on a  precision
     check data form  (Figure 3.3), and
     mark the chart as "unadjusted"
     precision check.

The  biweekly check  generates data for
assessing the precision of the
monitoring data; Section 2.0.8 of this
volume of the Handbook presents
procedures for calculating and
reporting precision.

3.4,4  Special Instructions for
Precision Checks on Beckman Model
866 Ambient Carbon Monoxide
 Analyzer — Because of the operational
 nature of the Beckman Model 866 CO
 analyzer, the following slightly
 modified procedures for precision
 checks and audits of this  analyzer
 model are generally necessary to
 obtain accurate quality assessment of
 the ambient readings.
   The Model  866 uses a dynamic,
 flowing reference cell  as part of its
 compensation for variable
 environmental factors such as water
 vapor and carbon dioxide  (COz). This
 mechanism responds rather slowly to
 changes in water vapor concentration.
 Although the syste.m is entirely
 adequate to follow natural
 environmental, moisture changes, it
 does not respond instantly to rapid
 changes in moisture level that occur
 when the analyzer is switched  from
 ambient air to dry concentration
 standards used for precision checks
 and audits. Most concentration
 standards obtained from compressed
 gas cylinders  or diluted from  high-
 concentration gas cylinders have a very
 low moisture  level, whereas ambient
 air normally contains much higher
 levels of water vapor. During the
 period immediately following a  switch
 from ambient sampling to a
 concentration standard, the analyzer
 is operating in a  nonequilibrated
 mode, which causes a significant
 offset (up to 1 to 2 ppm) in the
 analyzer's readings.  Accordingly, the
 precision check or audit response will
 be inaccurate unless suitable
 compensatory measures are taken.
 (This effect is accounted for in the
 calibration and automatic
 standardization procedures in the
 operation manual; accurate-calibration
 and .automatic standardization will  be
 obtained if these procedures are
 followed explicitly.)

  Either of two methods may be used
to avoid errors from this effect during
precision checks and audits. The first
is simply to allow sufficient time for
the analyzer to reestablish equilibrium
at the concentration-gas moisture
level. Equilibrium is established when
the analyzer response to this
concentration  standard stabilizes at  a
new reading somewhat different than
the original reading. (The original
reading may be stable for 10 to 20
minutes after introduction  of the dry
gas before the offset occurs.)
Unfortunately, the analyzer may
require as much as 1 to 2  hours to
reach moisture equilibrium at the dry-
gas condition.
  The second method takes advantage
of the temporary stable period
immediately after dry gas is
introduced, and it must be completed
before the offset occurs. Prior to the
precision check or audit, dry zero gas
is introduced into the analyzer just
long enough to establish a
temporary,  nonequilibrated zero
baseline. The precision or audit
concentration standard(s) is then
introduced, and interpretation of the
reading(s) is based on the net
response referenced to this temporary
zero baseline rather than the
equilibrated zero baseline; i.e., the net
difference between the response to
the standard and the temporary
baseline is  used with the calibration
curve to determine the response in
concentration units. Finally, dry zero
gas is reintroduced to verify that the
offset has not yet occurred and that
the temporary zero baseline has not
shifted. If the temporary zero baseline
has changed significantly, the second
method is not valid and the precision
check or audit must be repeated by
using the first method.

-------
                                 Jan. 1983
                                                                                     Section 2.6.3
Site ID	




Location  _




Address  /[7£>
                                                                 Pollutant
                                                                 Serial number
Date

I
Operator


i
Precision
test gas
concentration,
ppm
/O Of^i^\
i

Analyzer
response
% chart
"2—b

ppm


!
Difference.
ppm



Figure 3.3.   Example of precision check form.

-------
                         Section 2.6.3
                                                                           Jan. 1983
 Table 3.1.     Daily A ctivity Matrix
Characteristic
Acceptance limits
Frequency and method
of measurement	
Action if
requirements
are not met
Shelter temperature



Sample introduction system



Recorder




Analyzer operational settings




Analyzer operational check




Precision checks
Mean between 22° and
28°C (72° and 82°F); daily
fluctuations not greater
than ±2°C (4°F)
No moisture, foreign
materials, leaks, or  ob-
structions; sampling line
connected to manifold
Adequate supply of ink and
chart paper; legible  ink
traces; correct setting of
chart speed and range
switches; and correct time
Flow and regulator indica-
tors at proper settings:
analyzer in sampling mode;
and zero and span controls
locked at proper settings
Zero and span within toler-
ance limits (Subsec 9.1.3,
Sec. 2.0.9)
Precision assessed
(Subsec. 3.4.3)
Edit thermograph chart
daily for variations <2°C
(4°F)

Weekly visual inspection
Weekly visual inspection
Weekly visual inspection
Level 1 zero and span
checks every 2 weeks:
Level 2 checks between
Level 1 checks at frequency
decided by user
Every 2 weeks {Subsec.
3.4.3)	
Mark the strip chart for the
affected time period: repair
or adjust the temperature
control system.
Clean,  repair, or replace as
needed.
Replenish the ink and chart
paper supplies: adjust the
recorder time to agree with
clock, and note the time on
on the chart.
Adjust or repair as needed.
Isolate the source of error,
arid then repair; after cor-
rective action, recalibrate
the analyzer

Calculate and report the
precision (Sec. 2.0.8)	

-------
                              Jan. 1983
                                                                              Section 2.6.4
                        4.0   Data  Reduction,  Validation, and Reporting
    Quality assurance activities for data
  reduction, validation, and reporting
  are summarized in Table 4.1 at the
  end of this section.

  4.1   Data Reduction
    Hourly average concentrations from
  a strip chart record  may be obtained
  by the following procedure:
    1. Make sure the strip chart record
      has a zero-trace at the beginning
      and end of the sampling period.
   2.  Fill in the identification data
      called for at the top of the hourly
      average data form (Figure 4.1).
   3.  Draw a line from the zero
      baseline at the start of the
      sampling period to the zero
      baseline at the end of the
      sampling period by using a
     straight edge.
   4.  Read the zero baseline (% Chart)
     at the midpoint of each hourly
     interval, and record the value on
     the data form.
   5. Determine the hourly averages
     by placing a transparent straight
     edge parallel to  the horizontal
     chart division lines. Adjust the
     straight edge between the lowest
     and  highest points of the trace in
     the interval  between two vertical
     hour lines of interest so that the
     area above the  straight edge and
     bounded by the trace and the
     hour  lines is approximately equal
     to the area below the straight
     edge  and bounded by the trace
     and hour lines,  as shown below.
                         8. Convert reading values (% chart)
                            to concentrations (ppm) by using
                            the most recent calibration curve,
                            and record the CO
                            concentrations in the last column
                            of an hourly averages form such
                            as that shown in Figure 4.1.
                         An alternative method of converting
                       % chart to ppm is to eliminate steps 6,
                     •  7, and 8 and to use Equation 4-1:

                                             YV
                                   ,-,-...    - "i
                                         Slope
                                              Equation 4-1
                       where
                            Yz=zero baseline from step 4,
                                % scale
                             Y = recorder reading from step
                                5, % scale
                         Slope = slope of calibration
                                relationship from Section
                                2.6.2

                       4.2  Data Validation
                         Data of poor quality can be worse
                       than no data. Data validation to
                       screen for possible  errors or.
                       anomalies is one activity of a quality
                       assurance program. Statistical
                       screening procedures should be
                       applied to identify gross anomalies in
                       air quality data.6 Subsections 4.2.1
                       and  4.2.2 recommend two data
                       validation checks.

                       4.2.1  Span Drift Check — The first
                     •level of data validation for accepting
                       or rejecting the monitoring data
                       should be based on  routine periodic
   50
   4O
   30
   20
   JO
   0\
Straight edge
                            Area above line

                            Area below line
                1200
                             13OO
                                         J40O
                                                       150O
                                                                   1600
Read the deflection (% chart) for all of
the hourly intervals for which data
have not been marked invalid, and
record all values on the hourly
average data form in the column
headed Reading - Original (Orig).
  6. Subtract the zero baseline value
    from the reading value,  and
    record the difference.
  7. Add the percentage of zero
    offset,  <-5 percent, to each
    difference.
  checks of the analyzer. Results from
  the Level 1  span checks (Section
  2.6.3) should be used as the first
  Level of data validation. Thus, up to 2
  weeks of monitoring data may be
  invalidated if the span drift for a
  Level 1 span check is >25 percent."
  For this reason, it may be desirable to
  perform Level 1 checks more often
  than the recommended 2-week
  frequency.

  4.2.2  Edit of Strip Chart — The strip
  chart should be edited to detect signs
  of monitoring-system malfunctions
  that result in traces that do  not
  represent "real" data. In a review of a
•  strip chart, typical, points to watch for
  are:
   1. A straight trace (other than
      minimum detectable) for several
      hours.
   2. A wide solid trace indicating
      excessive noise or spikes that are
      sharper than  is possible with the
      normal instrument response time
      and are indicative of erratic
      behavior. Noisy outputs usually
      result when analyzers are
   -  exposed to vibration sources.
   3. A long, steady increase or
      decrease in deflection.
   4. A cyclic trace pattern within  a
      definite time period, which
      indicates a sensitivity to changes
      in temperature or parameters
     other than CO concentration.
   5. A trace that drops below the zero
     baseline during certain  periods;  •
     this may indicate a larger-than-
     - normal  drop in the ambient room
     temperature or the power line
     voltage. This may also indicate
     CO in the zero-air.
 Void any data for any time interval for
 which a malfunction of the sampling
 system is detected.

4.3   Data Reporting
  Information and data from the
hourly average form should be
transcribed to a SAROAD hourly data
form (see Section 2.0.3 of this volume
of the Handbook for details and
instructions for filling out the
SAROAD). If the data are to be placed
in the National Aerometric Data Bank,
further instructions  can be obtained
from the SAROAD Users Manual.7

-------
                           Section 2.6.4
                                                                                 Jan. 1983
         £)a\Hrw\  ,
 Calibration curve: Slope (bl *
Site number   3(?lbb£>C>/7
                                                               Pollutant
                                                                          OO
                                                                          B
Intercept (a) =
x = fy-aj/b
Date
b-15
fc-(5
















Hour
00
01






--









Reading
Orig
&l
zz
















Check









•








Zero baseline
Orig
t,
(,
















Check









•








Difference
Orig
/5
17
















Check


















y. Add + 5
Orig
2.0
zz
















Check


















x, ppm
Orig
8
7
















Check


















Flgut* 4.1.   Sample data form for recording hourly averages.

-------
                              Jan. 1983
                                                                              Section 2.6.4
 Table 4.1.    Activity Matrix for Data Reduction
Activity
                                Acceptance limits
                             Frequency and method
                             of measurement
                                                                                         Action if
                                                                                         requirements
Data reduction


Span drift check




Editing of strip chart


Data reporting
Stepwise procedure for
data reduction (Subsec.
4.1}
Level 1 span check <25%
(Sec. 2.6.3)
No sign of malfunction
Data transcribed to
SAROAD hourly data form
Follow method Subsec. 4.1
for each strip chart.

Perform Level 1 check at
least every 2 weeks. (Sec.
2.6.31
Visually edit each strip
chart. (Subsec. 4.2)

VisuaJ checks
 Review data reduction
 procedure.

. Invalidate data; take
 corrective action; increase
 frequency of Level 1
 checks until data are
 acceptable.
 Void data for time interval
 for which malfunction of •
 sampling system detected.
 Review data transcription

-------

-------
                             Jan. 1983                    1                 Section 2.6.5
                                           5.0   Maintenance
   The quality assurance activities for
 maintenance are described briefly.

 5.1   Preventive Maintenance
   Maintenance requirements vary
 from instrument to instrument;
 therefore, the supervisor should refer
 to the manufacturer's manual for a
 thorough discussion of maintenance
 requirements for a specific analyzer.
 After becoming familiar with the
 requirements, the supervisor should
 develop a suitable preventive
 maintenance schedule.

 5.2   Corrective Maintenance
  Corrective maintenance is any
 unscheduled maintenance activity that
 becomes necessary because of system
 malfunctions; for example,
 replacement of a damaged pump
 diaphragm, cleaning of a clogged
 sampling line, or replacement of a
 defective temperature control card.
 The need for corrective maintenance
 becomes apparent as the operator
 performs the daily operations
 described in Section 2.6.3 of this
 Handbook;"when the need arises, the
 operator should  refer to the
 manufacturer's manual for
 troubleshooting procedures.  A detailed
 maintenance record should  be kept on
file to identify recurring system
malfunctions. A  sample maintenance
log is presented  in Figure 5.1.

-------
                           Section 2.6.5
                                                                                 Jan. 1983
Site identification

Address
                                                              Pollutant
                                                               Instrument
                                                               Serial number
    Date
                   Initiajs of
                   technician
   fvent
  initiating
maintenance
maintenance
   activity
Comments
                                                                             Dai
Figure S. J,   Analyzer maintenance log.

-------
                              Jan. 1983
                                                                              Section 2.6.6
                                        6.0   Auditing  Procedure
   An audit is an independent
  assessment of the accuracy of data
  generated by an ambient air analyzer
  or a network of analyzers.
  Independence is achieved by having
  the audit performed by an operator
  other than the one conducting the
  routine field measurements and by
  using audit standards, reference
  materials, and equipment different
  from those routinely used in
  monitoring.
   The audit should be an assessment
  of the measurement process under
  normal operations-that is, without
  any special preparation or adjustment
  of the system. Routine quality
  assurance checks (e.g., those in
  Section 2.6.3) conducted by the
  operator are necessary for obtaining
 and reporting good quality data, but
 they are not to be considered part of
 the  auditing procedure.
   Three audits are recommended: two
 performance audits and a system audit.
 The performance audits are described
 in detail in Subsection 6.1, and the
 system aud.it is described in
 Subsection 6.2. These audit activities
 are summarized in Table 6.1 at the  •
 end of this section. (See Sections
 2.0.11 and 2.0.12 for detailed
 procedures for a system audit and a
 performance audit, respectively.)
   Proper implementation of an
 auditing program will ensure the
 integrity of the data and assess the
 accuracy of the data. The technique  '
 for estimating'the accuracy of the
 data is presented in Section 2.0.8 of
 this  volume  of the Handbook.


 6.1    Performance Audits
  The following subsections describe
 the recommended performance  audits:

 6.1.1  Calibration Audit —
  A calibration audit consists of
 challenging the continuous analyzer
 with known concentrations of CO
 within the measurement  range of  the
 analyzer. Known concentrations of CO
 can be generated  by using individual
 cylinders for each concentration or by
 using one cylinder of a high CO
 concentration and diluting it to the
desired levels with zero-air. In either
 case, the gases used must be
 traceable to an NBS-SRM or a
commercially available CRM (Section
 2.6.2); acceptable  protocol for
demonstrating traceability is
  presented in Section 2.0.7. A dynamic
  dilution system must be capable of
  measuring and controlling flow rates
  to within ±2 percent of the required
  flow. Flow meters must be calibrated
  under the conditions of use against a
  reliable standard such  as  a soap
  bubble meter or a wet  test meter; all
  volumetric flow rates should  be
  corrected to STP  at 25°C (77°F) and
.  760 mm (29.92 in.) Hg, but if both the
  CO and the zero air flow rates are
  measured with the same type device
  at the same temperature and
  pressure, the STP correction factor in
  the audit  equations can be
  disregarded. Note: If a wet test meter
  or a bubble meter is used for flow
  measurement, the vapor pressure of
  water at the temperature of the meter
  must be subtracted from the barometric
  pressure.
   The audit schedule depends on the
  purpose for which the monitoring data
  are being collected. For SLAMS monitor-
  ing,  each  analyzer must be audited at
  least once a year.  Each  agency should
  audit 25 percent of  the Reference or
  Equivalent analyzers each quarter.8 If
  an agency operates fewer'than  four
  analyzers, they should be randomly
 selected for  reauditing  so  that  one
 analyzer  is audited each  calendar
 quarter and each  analyzer audited at
 least once a year. For PSD monitoring,
 each Reference or Equivalent analyzer
 must be audited at least once during a
 sampling quarter.9

 6.1.2  Calibration A udit
 Procedures —
   The analyzer should be challenged
 with  at least one audit gas of known
 concentration from each of the
 following concentrations within the
 measurement range of the analyzer
being audited:
Audit point
1
2
3
4
CO concentration range.
ppm
3 to 8
15 to 2O
35 to 45
80 to 9O
The difference in CO concentration
(ppm) between the audit value and the
measured value is  used to calculate
the accuracy (Section 2.0.8) of the
analyzer.
  Information on the station,
analyzer, audit device, reference
  materials, and audit procedures are of
  prime importance because the validity
  of the audit results depends on
  accurate documentation (Figures 6.1
  and 6.2). The following procedure has
  been, developed to aid in conducting
  the audit.
    1. Record the station's number,
  name, and address on the audit
  summary report (Figure 6.1).
   2. Identify the perso.n(s)
  performing the audit and indicate the
  date of the audit.
   3. Record the type of audit device
  used. If it was purchased,  record the
  manufacturer's name, model, and
  serial number; if it was assembled by
  the .user, assign an identification
  number so that audits can be
  referenced to that particular
  apparatus.
   4. Identify the CO cylinder(s) used
  for auditing and the NBS-SRM or
  commercially available CRM  used to
  verify the concentration. As required,
  the CO cylinder(s) should be
  reanalyzed every 6 months (Section
  2.0.7).
   5.  Identify the device used to
 measure flow rates, if applicable;
   6. "Connect the  audit system outlet
 line to the inlet of the CO analyzer.
 Analyzers  must operate in  the normal
 sampling mode during the  audit, and
 the test atmosphere must pass
 through all filters, scrubbers,
 conditioners, and other components
 used during normal  ambient sampling
 and through as much of the ambient
 air inlet system as practicable. The
 exception to this rule that is permitted
 for certain CO  analyzers during
 precision and span checks does not
 apply for audits.

   7. Turn on the zero-air flow, and be
 sure that the zero  air output exceeds
 the analyzer intake by at-least 10
 percent.
   8. Record the analyzer zero value on
 the audit summary report.
  9.  Generate the  first up-scale audit
point by challenging  the  analyzer with a
CO concentration within one of the
required concentration ranges; obtain a
stable trace, and record the audit value
and the analyzer response on  the audit
summary report.
  10.  Determine the analyzer's
response (ppm) from  the analyzer's
latest calibration relationship;  if the
relationship is reported as slope and

-------
                           Section 2.6.6
                                                                                Jan. 1983
   Station
                                                             2, Analyzer mfr. .

                                                                      Model
 3. Audit performed by

 4. Audit device mfr. .

            S/N 	
                                                              Date
                                                              Model
 5,  CO standard used
   Verified' a/jafnst NBS-SRM

   By .
                            rn.  c
-------
                               Jan. 1983
                                                                                Section 2.6.6
    Yes     No
                    1. Zero and span checks performed at least biweekly

                    2. Temperature variations monitored


                    3. Flow meters routinely calibrated to ±2% accuracy against a reliable standard such as a soap bubble meter or
                      wet test meter

                    4. Flow rates monitored routinely

                    5. Excessive noise minimized


                    6. Data processing checks performed


                    7. Multipoint calibration performed routinely, and results of the calibrations recorded


                    8. Quality control charts maintained for zero and span checks


                    9. Maintenance performed routinely on pertinent components per manufacturer's manual

                   10. Calibration gases traceable to an NBS-SFtM


                   1J. Sample introduction system check made weekly


                   12. Paniculate filter (if used) changed per manufacturer's manual


                   13. Recording system checked and serviced before each sampling period

                   14. Recorded data checked for signs of system malfunction

                   15. Data quality records maintained — completeness, accuracy, precision, and representativeness

                   16. Calibration gases periodically assayed against an NBS-SRM

                  Comments: 	:	:	
 Figure 6.2.    Checklist for use by auditor. (Measurement of Continuous CO in Ambient Airj
 intercept, use Equation 6-1 of Fiqure
 6.1.
   11.  Repeat steps 9 and  10 for two
 more audit points.
   12.  Calculate the percent difference
 for each audit point by using  Equations
 6-2 and 6-3 of Figure 6.1 and record on
 the audit summary report.  Results of
 the audit are used to estimate the
 accuracy of ambient air'quality data (as
 described in Section 2.0.8).

 6.1.3   Data  Reduction Audit — Data
 reduction involves reading  a  strip
 chart record,  calculating an average,
 and either transcribing or recording
 the results on the SAROAD form. The
 audit is an independent check of the
 entire data reduction process, and
 should be performed by an individual
 other than the one who originally
 reduced  the data. Initially, the data
processing check should be performed
for 1 day out  of every 2 weeks of data.
For two 1-hour periods within each
day audited, make independent
 readings of the strip chart record and
 continue through the actual
 transcription of the data on the
 SAROAD form. The 2 hours selected
 during each day should be those for
 which the trace is either most
 dynamic (in terms of spikes), or for
 which the  average concentration is the
 highest.
   The data processing check is made
 by calculating the difference:

           d = [CO]B - [COJA
                         Equation 6-4
 where
      d = the difference between the
         measured value and the
         corresponding check value,
         ppm
   [CO]R=the recorded analyzer
         response, ppm
   [CO]A = the audit value of the CO
         concentration, ppm
   If d exceeds ±2 ppm, all of the
remaining data in the 2-week period
should be checked.
 6.2  System Audit
   A system audit is an onsite
 inspection and review of the quality
 assurance activities used for the total
 measurement system (sample
 collection, sample analysis, data
 processing, etc.). System audits are
 normally qualitative appraisals of
 system quality conducted at the
 startup of a new monitoring system
 and periodically,  as appropriate, to
 audit significant changes in system
 operation.
  An example form for a system audit
 is shown in Figure  6.2. The items on
 this form should  be checked for
 applicability to  the particular local.
 State, or Federal  agency.
  See Sections 2.0.11 and 2.0.12 for
detailed procedures and forms for a
system audit and a  performance audit.
respectively.

-------
                         Section 2.6.6
                                                                           Jan. 1983
 Table 6,1.    Activity Matrix for A udit Procedure
Audit
Acceptance limits
Frequency of method
of measurement	
Action if
requirements
are not met
Multipoint calibration audit
Data processing audit
System audit
The difference in concen-
trations between the
measured values and the
audit values is used as a
measure of accuracy.
(Sec. 2.0.8)
Adhere to stepwise
procedure for data reduc-
tion. Sec. 2.6.4; no
difference should exceed
±2 ppm.

Use method described in
this section of the  Hand-
book.
Perform at least once per
quarter; see Subsec. 6.1.1
for procedure.
Perform independent data
processing check on a
sample of the recorded
data; e.g., check 1 day out
of every 2 weeks of data. 2
hours for each day.
Perform at the start-up
of a new monitoring sys-
tem, and periodically as
appropriate; observation
and checklist IFiq. 6.21.
If differences are outside
the agency acceptance
limits, locate the problem
and correct.
Check all remaining data
if one or more data reduc-
tion checks exceed ±2
ppm.
Initiate improved methods
 and/or training programs.

-------
                           Jan. 1983                   1                 Section Z.6.7
           7.0   Assessment of Monitoring Data for Precision and Accuracy
  For continuous analyzers, perform a
check every two weeks to assess the
precision of the data. Use these data
to estimate single instrument
precision as described in Section
2.0.8 of this volume of the Handbook.
The precision check procedures
described in Section 2.6.3 are
consistent with those given in
References 8 and 9.
  Estimates of single instrument
accuracy for ambient air quality
measurements from continuous
methods are calculated according to
the procedure in Section 2.0.8. The
audit procedure is described in
Section 2.6.6.

-------

-------
                             Jan. 1983
                                                                           Section 2.6.3
                8.0   Recommended Standards for Establishing Traceability
    Two considerations are essential for
  ensuring data of the desired quality:
    1. The measurement process must
      be in statistical control at the
      time of the measurement.
    2. The systematic errors, when
      combined with the random
      variation in the measurement
      process, must result in an
      acceptable uncertainty.
    Evidence of good quality data
  includes documentation of the quality
  control checks and the independent
  audits of the measurement process by
  the recording of data on specific forms
  or on a quality control chart and by
  using materials, instruments and
  measurement procedures that can be
 traced to appropriate standards of
 reference.
   For traceability to be established,
. data must be obtained routinely by
 repeat measurements of standard
 reference samples (primary.
secondary, and/or working standards).
A condition of process control also
must be established. Working
calibration standards should be
traceable to standards of higher
accuracy.
  The CO calibration standards must
be traceable to an NBS-SRM (as listed
in Table 8.1) or to a commercially
available CRM.
  A list of gas manufacturers who
produce approved CRM is available by
writing to:
  U.S. Environmental Protection
  Agency
  Environmental  Monitoring Systems
  Laboratory (MD-77)
  Research Triangle Park, North
  Carolina 27711  -•
  ATTN: List of CRM  Manufacturers
Tables. 1.    NBS-SRM's for CO Monitors
SRM
168O
1681
2613
2614
Type
CO in nitrogen
CO in nitrogen
CO in air
CO in air
Vol/unit,
liters
at STP
870
870
870
870
Nominal CO
concentration,
	 ppm 	
500
7000
18. 1
43 0

-------

-------
                               Jan. 1983
                                                                              Section 2.6.3
                                        9.0   Reference Methods*
   Appendix C—Measurement
   Principle and Calibration
   Procedure for the
   Measurement of Carbon
   Monoxide in the Atmosphere
   (Non-Dispersive Infrared
   Photometry)

   Measurement Principle

   1. Measurements are based on the
   absorption of infrared radiation by
   carbon monoxide (CO) in a non-
   dispersive photometer. Infrared
   energy from a source is passed
   through a cell containing the gas
   sample to be analyzed, and the
  quantitative absorption of energy by
   CO in the sample cell is measured
  by .a suitable detector. The
  photometer is sensitized to CO by
  employing CO gas in either the
  detector or in a filter cell in the
  optical path, thereby limiting  the
  measured absorption to one or more
  of the characteristic  wavelengths at
  which CO strongly absorbs. Optical
  filters or otfier means may also be
  used to limit sensitivity of the
_ photometer to a narrow band of
  interest. Various schemes may be
  used to provide a  suitable zero
  reference for the photometer. The
  measured absorption is converted to
  an electrical output signal, which-is
  related to the concentration of CO
  in the measurement  cell.
  2. An analyzer based on this
  principle will be considered a
  reference method only if it has been
  designated as a reference method in
  accordance with Part 53 of this
  chapter.

  3.  Sampling considerations.
   The use of a particle filter on the
  sample inlet line of an NDIR CO
  analyzer is optional and left to the
  discretion of the user or the
  manufacturer. Use of filter should
  depend on the analyzer's'
  susceptibility to interference,
  malfunction, or damage due to
 particles.

  Calibration  Procedure

 1.  Principle. Either of two methods
 may be used for dynamic
   multipoint calibration of CO
   analyzers:  (1) One method uses a
   single certified standard cylinder of
   CO, diluted as necessary with zero
   air, to obtain the various calibration
   concentrations needed. (2) The other
   method uses individual certified
   standard cylinders of CO for each
   concentration needed. Additional
   information on calibration may be
   found in Section 2.0.9 of Reference
   1.

   2.  Apparatus. The major
  components and typical
  configurations  of the calibration
  systems for the two calibration
  methods are shown in Figures 1 and
 : 2.

  2.1  Flow controller(s). Device
  capable of adjusting and regulating
  flow rates. Flow rates for the
  dilution method (Figure  1) must be
  regulated to ±/%.

  2.2  Flow meter(s). Calibrated flow
  meter capable of measuring and
  monitoring flow rates. Flow rates for
  the dilution  method (Figure 1) must be
  measured with an accuracy of ± 2%
  of the measured value.

  2.3  Pressure regulator(s) for
  standard CO cylinder(s). Regulator
  must have nonreactive  diaphragm and
  internal parts and a suitable delivery
  pressure.
 2.4  Mixing chamber. A chamber
 designed to provide thorough mixing
 of CO and diluent air for the
 dilution method.

 2.5  Output manifold. The output
 manifold should be of sufficient
 diameter to insure an insignificant
 pressure drop at the analyzer
 connection. The system must have a
 vent designed to insure atmospheric
 pressure at the  manifold and to
 prevent ambient air from entering the
 manifold.

 3.  Reagents.

 3.1  CO concentration standard(s).
   Cylinder(s) of CO in air containing
 appropriate cbncentration(s) of CO
 suitable for the selected operating
 range of the analyzer  under
'calibration; CO standards for the
 dilution method may be contained in a
 nitrogen matrix if the  zero air dilution
 ratio is  not less than 100:1. The assay
 of the cylinder(s) must be traceable
 either to a National Bureau of
 Standards (NBS) CO in air Standard
 Reference Material (SRM) or to  an
 NBS/EPA-approved commercially
 available Certified Reference Material
(CRM). CRM's are described in
 Reference 2, and a list of CRM
sources is available from the address
shown for Reference 2. A
recommended protocol for certifying
                                                    Mixing
                                                   Chamber
                                                CO
  CO
  Std
 Output
 Manifold
                 Vent
                                                     P
                  Extra Outlets Capped
                    When Not in Use
      To Inlet of Analyzer
       Under Calibration
•40 CFR 50 Appendix C (as amended 47 FR
54922. December 6. 1982)
Figure 1.    Dilution method for calibration of CO analyzers.

-------
                        Section 2.6.9
                                                                           Jan. 1983
fS t-1-*- Flow riowmetei 	 1
i i i ^ ^ ^ C°ntr°"er "*"'"''*
A
CO
Sid



n
CO
Std



A
CO
Stf



Ve
A
CO
Sfrf



A
Zero
Air



Output
Manifold
di tt III
t
                        Extra Outlets Capped
                          When Not in Use
            Jo Inlet of Analyzer
             Under Calibration
Flguti 2,   Multiple cylinder method for calibration of CO analyzers.
CO gas cylinders against either a CO
SRM or a CRM is given in Reference
1, CO gas cylinders should be
recertified on a regular basis as
determined by the local quality control
program,

3.2  Dilution gas (zero air). Air, free
of contaminants which will cause a
detectable response on the CO
analyzer. The zero air should contain
<0.1 ppm CO. A procedure for
generating zero air is given in
Reference 1.

4.  Procedure Using Dynamic
Dilution Method.

4.1  Assemble a dynamic calibration
system such as the one shown in
Figure 1, All calibration gases
including zero air must be introduced
into the sample inlet of the analyzer
system. For specific operating
instructions refer to the
manufacturer's manual.

4.2  insure that all flowmeters are
properly calibrated, under the
conditions of use. if appropriate,
against an authoritative standard
such as a soap-bubble meter or wet-
test meter, All volumetric flowrates
should be corrected to 25°C and 760
mm Hg (101 kPa). A discussion on
calibration of flowmeters is given in
Reference 1;

4.3  Select the operating range of
the CO analyzer to be calibrated.

4.4  Connect the signal output of the
CO analyzer to the input of the strip
chart recorder or other data collection
device. All adjustments to the
analyzer should be based on the
appropriate strip chart or data device
readings. References to analyzer
responses in the procedure given
below refer to recorder or data device
responses.

4.5  Adjust the calibration system to
deliver zero air to the output manifold.
The total air flow must exceed the
total demand of the analyzer(s)
connected to the output manifold to
insure  that no ambient air is pulled
into the* manifold vent. Allow the
analyzer to sample zero air until a
stable response is obtained. After the
response has stabilized, adjust the
analyzer zero control. Offsetting the
analyzer zero adjustments to + 5
percent of scale is recommended to
facilitate observing negative zero drift.
Record the stable zero  air response as
Zco.

4.6  Adjust the zero air flow and the
CO flow from the standard CO
cylinder to provide a diluted CO
concentration of approximately 80
percent of the upper range  limit (URL)
of the operating range  of the analyzer.
The total air flow must exceed the
total demand of the analyzer(s)
connected to the output manifold to
insure  that no ambient air is pulled
into the manifold vent. The exact CO
concentration is calculated  from:
        [COlour = [COlsToxFco
                   FD + FD


Where:
  [CO]ouT = diluted CO concentration
          at the output manifold,
          ppm;
(D
  [CO]sTo = concentration of the
           undiluted CO standard,
           ppm;
      FCo = flow rate of the CO
           standard corrected to 25°C
           and 760 mm Hg, (101 kPa),
           L/min; and
       Fo = flow rate of the dilution
           air corrected to 25°C and
           760 mm Hg, (101  kPa),
           L/min.
  Sample this CO concentration until a
stable response is obtained. Adjust
the analyzer span control to obtain a
recorder response as indicated below:

Recorder response (percent scale) =
         [CO]ouTx100 + Zco
           URL
                                 (2)
Where:
  URL = nominal upper range limit of
        the analyzer's operating range,
        and
    Zco = analyzer response to  zero air,
       ' % scale.
  If substantial adjustment of the
analyzer span control is required, it
may be  necessary to recheck the zero
and span adjustments by repeating
Steps 4.5 and 4.6. Record the CO
concentration and the analyzer's
response.

4.7  Generate several additional con-
centrations (at least three evenly spaced
points across the remaining scale are
suggested to verify linearity) by decreas-
ing  Fco or  increasing FD.  Be  sure the
total flow exceeds the  analyzer's total
flow demand. For each concentration
generated,  calculate the exact CO
concentration using Equation (1).
Record the  concentration  and the
analyzer's response for each concentra-
tion. Plot the analyzer responses versus
the  corresponding CO concentrations
and  draw or calculate the calibration
curve.

5.   Procedure Using Multiple
Cylinder  Method.
  Use the procedure for the dynamic
dilution method with the following
changes:

5.1   Use a multi-cylinder system
such as the typical one shown in
Figure 2.

5.2  The flow meter need not be
accurately calibrated, provided the
flow in the  output manifold exceeds
the  analyzer's flow demand.

5.3  The various CO calibration
concentration required in Steps 4.6 and
4.7  are obtained without dilution by
selecting the  appropriate certified
standard cylinder.

-------
                                                                           Section 2.6.9
                            Jan. 1983
 References
   1. Quality Assurance Handbook for
 Air Pollution Measurement Systems.
 Volume II—Ambient Air Specific
 Methods, EPA-600/4-77-027a, U.S.
 Environmental Protection Agency,
 Environmental Monitoring Systems
 Laboratory, Research Triangle Park,
 North Carolina 27711, 1977.
  2. A Procedure for Establishing
 Traceability of Gas Mixtures to Certain
 National  Bureau of Standards
 Standard 'Reference  Materials. EPA-
 600/7-81-010, U.S. Environmental
 Protection Agency, Environmental
Monitoring Systems  Laboratory (MD-
77), Research Triangle Park, North
Carolina 27711, January 1981.

-------

-------
                             Jan-1983                    1                Section 2.6.10
                                          10.0   References
   1. 40 CFR 50.8.
   2. 40 CFR 53.
   3. 40 CFR 50, Appendix C. (As
 amended 47 FR 54922, December 6
 1982).
   4. U.S. Department of Commerce.
 Catalog of NBS Standard Reference
 Materials. NBS Special Publication
 260, 1981-1983 Edition. National
 Bureau  of Standards, Washington,  D.C.
 November 1981.
   5. U.S. Environmental Protection
 Agency. Field Operations Guide for
 Automatic Air Monitoring Equipment.
 Office of Air Programs. Publication
 Nos. APTD-0736, PB 202-249 and PB
 204-650, October 1972.
  6.  U.S. Environmental Protection
Agency.  Screening Procedures for
Ambient Air Quality Data. EPA-
450/2-78-037, July 1978.
  7.  U.S. Environmental Protection
Agency.  Aeros Manual Series Volume-
II: Aeros User's Manual. EPA-450/2-
76-029,  OAQPS No. 1.2-039,
December 1976.
  8. 40 CFR  58, Appendix A.
  9. 40 CFR  58, Appendix B.   '

-------
4ft

-------

-------

-------
                                              Section No.  2.8
                                              Revision No.  0
                                              Date December 30,  1981
                                              Page 1 of  5
                            Section 2.8
         REFERENCE METHOD FOR THE DETERMINATION OF LEAD IN
            SUSPENDED PARTICULATE .MATTER COLLECTED FROM
                            AMBIENT AIR
                              OUTLINE
          0^+- •                                         Number of
          bec-cion                       Documentation     pages
SUMMARY                                      oo
                                             £. . O     .        J_
METHOD HIGHLIGHTS                    '        2.8             3
METHOD DESCRIPTION
    '• 1.   PROCUREMENT OF EQUIPMENT-      '
          AND SUPPLIES                       2.8.1           4
     2.   CALIBRATION OF EQUIPMENT      '     2.8.2           4
     3.   FILTER SELECTION AND
          PREPARATION                       .2.8.3           1
     4.   SAMPLING PROCEDURE                 2.8.4           1
     5.   ANALYSIS OF SAMPLES                2.8.5          14
     6.   CALCULATIONS AND DATA REPORTING    2.8.6           6
     7.   MAINTENANCE                        2.8.7         '  3'"
     8.   AUDITING PROCEDURE     '         "2.8.8          12
     9.   ASSESSMENT OF MONITORING DATA
          FOR PRECISION AND ACCURACY         2.8.9          \
    10.   RECOMMENDED STANDARDS FOR
          ESTABLISHING TRACEABILITY         2.8.10          1
    11.    REFERENCE METHOD                  2.8.11          6
    12.    REFERENCES                        2.8.12         2
    13.    DATA FORMS                        2.8^3

-------
                                             Section No. 2.8
                                             Revision No. 0
                                             Date December 30, 1981
                                             Page 2 of 5
                             SUMMARY

     Ambient  air  is  drawn through  a  glass  fiber  filter  of  a
hi-vol sampler  to collect particulate material.  The lead con-
centration  in the suspended  particulate matter is  analyzed by
atomic absorption spectrophotometry.1
     This method of sampling is applicable to measurement of the
mass  concentration of  suspended particulates  in  ambient air.
The  size  of the sample collected is usually adequate for other
analyses.  When the sampler  is operated 24 h at an average flow
rate  of  1.70 m3/min  (60.0 ft3/min) an  adequate sample  is ob-
tained,  even in  an  atmosphere having  a concentration  of sus-
pended particulates as low as 1 pg/m3.
     The  typical  range of the method is  0.07  to  7.5  pg Pb/m3
assuming  an  upper linear  range of  analysis  of  15  pg/ml and an
air volume of 2400 m3.  Typical sensitivities for a 1% change in
absorption  (0.0044  absorbance units) are  0.2 and  0.5  pg Pb/ml
for the 217.0 and 283.3 nm lines,  respectively.   A typical lower
detectable limit  (LDL)  is 0.07 pg  Pb/m3 when an air volume of
2400  m3  is  assumed.   The value quoted in  the Federal Register1
(0.07 pg Pb/m3) was derived from instruments of different models
using the 283.3 and the 217.0 nm Pb lines.
     Absolute values for  individual laboratories will vary with
the  type  of instrument, the  absorption  wavelength,  and the in-
strumental operating conditions.
     The  method description which  follows is based on the pro-
mulgated Reference Method.1

-------
                                              Section No.  2.8
                                              Revision No.  0
                                              Date  December 30,  1981
                                              Page  3  of 5
                        METHOD HIGHLIGHTS

     This  document is designed  as  a guideline for the develop-
ment  of  a quality  assurance  (QA)  program  as applied  to the
reference  method  for measurement of  ambient air particulate lead
concentrations.   The  description  of  the  lead method  is sub-
divided  into  categories,  for easy referral to specific informa-
tion.   Also  at  the  end of  each section is  an activity matrix
which  summarizes  the QA activities  and  can be used for a quick
review  of  important  information.   Any  of  the  sections,  data
forms, or  activity matrices  can be  easily removed  from the rest
of  the  Handbook  for making "working"  copies  for. convenient
reference.  The  sampling procedure  is  the same  as the ambient
air hi"-vol method  for  total particulate  described in Section 2.2
of Volume  II  of this Handbook.  Therefore, for complete sampling
and analytical procedures for ambient air particulate lead it is
necessary  to   refer  to  the  appropriate  Section/Subsections  of
Section  2.2.   Following is  a brief discussion of  the material
covered in the method  description.
1-   Procurement of Equipment and Supplies
     Section  2.8.1  describes the   equipment  and  supplies  for
monitoring ambient air particulate lead.  It includes specifica-
tions,  procurement  log,  acceptance  limits,  and  recommended
action  if  acceptance limits  are  not   met.   Table  1.1   is  an
activity  matrix   summarizing  the procurement of equipment and
supplies.
2.   Calibration of Equipment
     Section  2.8.2  provides  calibration  procedures  for  the
hi-vol sampler,  elapsed-time meter,  orifice  calibration  unit,
and the  atomic absorption spectrophotometer.   Table 2.1  is  an
activity  matrix   which summarizes  the   requirements  for  cali-
bration of the equipment.

-------
                                             Section No.  2.8
                                             Revision No. 0
                                             Date December 30,  1981
                                             Page 4 of 5            •

3.   Filter Selection and Preparation
     Section 2.8.3 briefly discusses filter selection.and prepa-
ration and  refers the reader  to Section 2.2.3  of  Volume  II  of
this Handbook  where the hi-vol  method  was  previously discussed
in detail.
4.   Sampling Procedure
     Section 2."8.4  refers  the  reader to Section 2.2.4 of Volume
II  of  this Handbook where the  hi-vol  sampling method  is dis-
cussed  in  detail.    Sampling  procedures 'for  total  suspended
particulate  and  ambient  air  particulate  lead are  identical.
5.   Analysis of Samples
     Section  2.8.5   contains  a  step-by-step  procedure  for  the
analysis of ambient hi-vol filters for lead.  The quality assur-
ance program  includes  sample  documentation and  inspection;  it
discusses   analytical   equipment   and  supplies,   analytical
reagents,  and calibration  of  the  analytical  procedure.  Table
5.1  is  an  activity matrix for  the analysis  and  it summarizes
acceptance limits of specific  activities involved with analysis
of samples.
6.   Calculations and Data Reporting
     Section 2.8.6 describes those.activities pertaining to data
calculation and  reporting.  An important part and final step in
this quality  assurance program  is  the  data calculation, review-
and  standardized reporting format.   Independent checks of the
data are  recommended any time  they are recorded or transcribed.
Example  data  forms   are  provided and an activity  matrix at the
end  of  this section summarizes  the activities for calculations
and data reporting.
7.   Maintenance
     Section  2.8.7   discusses  periodic  maintenance recommended
for  sampling  and analytical  equipment.   In order  to maintain
optimum equipment performance,  a preventive maintenance  schedule

-------
                                              Section  No.  2.8
                                              Revision No.  0
                                              Date December 30,  1981
                                              Page 5 of  5

 should  be  established.   Table  7.1  is  an activity  matrix  for
 maintenance  of  sampling  and analytical equipment.
 8.   Auditing Procedure
     Section 2.8.8 describes  the  recommended  audit procedures
 for  maintaining a  quality assurance  program- for  this method.
 There are two types of audits discussed:
     1.   performance audit
     2.   system audit.
 The  data  can be quantitatively evaluated by  performance audits.
 A  system  audit  will assess the quality of the total measurement
 system.  Example control charts and control limits are provided.
 Table 8.1 is an activity matrix which summarizes auditing proce-
 dures .
 9.  Precision and Accuracy
     Section 2.8.9  discusses  data evaluation based on precision
 and  accuracy.
 10.  Recommended Standards
     Section  2.8.10 discusses  traceability  of standards.  .As
 evidence  in  support of  good quality data,   it  is  necessary to
 perform quality control checks  and  independent audits  of  the
 measurement  process,  to  document  these  data,   and  to  use
 materials, instruments,  and  measurement procedures  that can be
 traced  to  appropriate   standards  of  reference.  This  section
 describes these  standards  of reference which are  applicable to
 this method.
     Section 2.8.11  is  the reference method  for measurement of
particulate lead in ambient air as promulgated by the EPA.
     Section 2.8.12 lists the references.
     Section  2.8.13  contains data  forms  applicable  to  this
method.   They can be easily removed and copied for use with this
method.

-------

-------
                                              Section No. 2.8.1
                                              Revision No. 0
                                              Date December 30, 1981
                                              Page 1 of 4
                        METHOD DESCRIPTION

 1.0  PROCUREMENT OF EQUIPMENT AND SUPPLIES
      Specifications for the equipment and  supplies  for monitor-
 ing ambient air for particulate lead (APPb) are  in the Reference
 Method,  Section 2.8.11  and listed below.
      Upon receipt  of the  sampling and the  analytical  equipment
 and supplies,  the procurement checks  should be conducted.   These
 quality  assurance checks are summarized  in Table  1.1  at the  end
 of  this  section.  These  checks  should be recorded in  a  procure-
 ment log.   An  example of a log  is  Figure  1.1 and a blank copy of
 the log  is  in  Section 2.8.13  for the  Handbook user.  The  procure-
 ment^, log will  serve  as a  permanent  record for future  procure-
 ments, provide continuity  of  equipment and  supplies, and provide
 records  for future program fiscal  projections.
     The  following is a  tabulation of equipment  and  supplies  and
 a  reference  to  the  subsections  within  this section  or   other
 sections  of  Volume II  of this  Handbook  containing  pertinent
 information.   This will  aid the user in  finding specific infor-
mation  regarding  equipment  and  supplies   necessary   for  this
method.
                                   Reference to Handbook
Equipment and Supplies   '     Section             Subsection
Sampling
 Hi-vol air sampler            2.2.7               7.1-7 6
Analysis equipment
 Atomic absorption spectro-
  photometer                   2.8.5               5.3.1
Analytical support
 Acetylene              •       2.8.5               5.3.2
 Air                           285               coo
                               •£•0-0               5.3.3

-------
                                             Section No.  2.8.1
                                             Revision No.  0
                                             Date December 30, 1961
                                             Page 2  of 4
I  C
o o
CL'i-
(/> 4->
  W
  O
  O
0)
4J
  £-
  o
  TD

  0)
  C
  (O


  01
  s_
  u
  U)
  O)
  -o

  e
  01
     s
     t
     o
     03
     00


     6
     I
     M
     M
h
O
                                                                 O)
                                                                 o
                                                            c
                                                            a>

                                                            
-------
                                              Section No.  2.8.1
                                              Revision No.  0
                                              Date  December 30,  1981
                                              Page  3  of 4
                                   Reference  to Handbook
Equipment and Supplies
Analytical support
 Beakers
 Volumetric flasks
 Pipettes
 Hot plate
 Ultrasonic bath
 Template
 Pizza cutter
 Watchglass
 Polyethylene bottles
 Parafilm "M"
Reagents
 Cone HNO3
 Cone HCl
 Water
 3M HN03
 0.45M HNO3
 2.6M HN03  +  0  to 0.9M HCl
 0.40M HN03 + XM HCl
 Pb(N03)2
 Lead stock solution
  (1000 |jg Pb/ml)
Section
2.8.5
2.8.5
.2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
2.8.5
Subsection
5.3.4
5.3.4
. 5.3.4
5.3.5
5.3.6 -
5.3.7
5.3.8-
5.3.9 .
5.3.10
5.3.11
5.4.1
5.4.2
' • 5.4.3
' ' 5.4.4
5.4.6
5.4.7
5.4.8'
• 5.4.9
5.4.10, 5.5.11

-------
                                                      Section No.  2.8.1
                                                      Revision No.  0
                                                      Date  December 30,
                                                      Page  4 of 4
                                                          1981
     TABLE 1.1.   ACTIVITY  MATRIX  FOR PROCUREMENT OF EQUIPMENT AND SUPPLIES
Equipment
Atomic absorp-
 tion spectro-
 photometer
Elapsed-time
 meter
Orifice cali-
 bration unit
Sampler
Acceptance limits
Equipped with lead hol-
low cathode lamp or
electrode!ess dis-
charge lamp
24 h ±2 min
Flow rate from manu-
facturer = actual  flow
rate ±4%
Sampler complete;
no damage evident
Frequency and method
   of measurements
On receipt,  check
specifications and
parts
On receipt,  check
against standard
timepiece of known
accuracy
On receipt, check
against flow rate
primary standard
Visually observe
Action if
requirements
are not met
Contact manu-
facturer or
supplier
Adjust or
replace
Adopt new
calibration
curve if no
evidence of
damage; re-
pi ace if
damaged
Repair or
replace

-------
                                              Section No. 2.8.2
                                              Revision No. 0
                                              Date December 30, 1981
                                              Page 1 of 4
 2.0  CALIBRATION OF EQUIPMENT
      Calibration of  the  equipment is one  of the most  important
 functions  for  maintaining  good  data  quality,  therefore,   the
 sampling and  analytical  equipment must be calibrated regularly.
      The calibration activities  are  summarized in Table 2.1,  at
 the end of this  section.   Many of these  activities should be  used
 as initial  procurement checks.   All  data and calculations  in-
 volved in the calibrations should be recorded in a calibration
 log,  which  should  have  a separate section  designated  for  each
 apparatus and  sampler.
 2.1  Hi-Vol  Sampler

      The hi-vol  air  sampler  shall be  calibrated as specified  in
 Section 2.2.2  of Volume II  of this Handbook.
 2'1-1   Elapsed-Time Meter -  The  elapsed-time' meter (synchronous.
 motor  type,  60 Hz)  should be  checked upon receipt and again every
 6  mo  against a timepiece of known  accuracy  (Section 2.2.2).
 2.1.2   On-off  Timer -  For  those  samplers  that are equipped with
 an on-off  timer,   the  timer   should be  calibrated  and  adjusted
 using  a calibrated elapsed-time meter,  upon  receipt and every.3
 mo  thereafter.   The  calibration  procedure is  in  Section  2.2.2.
 2'1-3   Orifice Calibration Unit - "The  orifice  calibration  unit
 should  be   calibrated  against  a  secondary   standard   (e.g.,  a
 Rootsmeter), upon  receipt  and at 1-year  intervals  thereafter.
 The calibration procedure is in Section 2.2.2.
 2.1.4  Sampler -   Samplers  must be  calibrated  when first  pur-
 chased,  after  major  maintenance  (e.g.,  replacement of  motor or
motor  brushes),  any time the  flow rate measuring device  (i.e.,
 rotameter  or recorder)  has  to be  replaced or  repaired, or  any
 time  a  one-point audit check  (Section 2.2.8)  deviates more  than
±7% from  the calibration  curve.   The  calibration procedure  for a
sampler is in Section 2.2.2.

-------
                                             Section No. 2.8.2
                                             Revision No. 0
                                             Date December 30, 1981
                                             Page 2 of 4
2.2  Atomic Absorption Spectrophotometer
     Major repairs and  adjustment  of the atomic absorption spec-
trophotometer normally  require  the  services  of the manufacturer
or the manufacturer's representative.   However,  before operating
the spectrophotometer,  instrument performance, such as sensitivi-
ty  and reproducibility  should be  checked by  using  a  standard
metal  solution.   The instrument  sensitivity depends  on- various
factors such  as  alignment of the hollow cathode  lamp and burner
head, cleanness of optical systems and. burner head, and the level
of grating system calibration.  An optimal sensitivity is recom-
mended by the manufacturer  in the instrument manual.  After this
sensitivity  is   achieved,   check  the  reproducibility.   Typical
sensitivities for a  1% change  in absorption are  0.2  and 0.5 jjg
Pb/ml  for wavelengths  ,of 217.0 and 283.3  nm,  respectively.   An
instrument reproducibility  change  of  ±5% is  acceptable;  repeat
the analysis until this limit is achieved.   If the reproducibili-
ty varies by more than ±5%,  the instrument should be checked by a
manufacturer's representative or a. qualified operator.
     The following step-by-step reproducibility test procedure is
recomended:
      1.  Prepare a  series  of standard lead solutions containing
0.2, 1.6, and 10.0   |jg  Pb/ml,- or 3  standard  concentrations that
"bracket1! the normal sample concentration  range,  as described in
Section 2.8.5, Subsection 5.7.2.
      2.  Set the  atomic absorption spectrophotometer for stan-
dard conditions,  as  described  in  Section  2.8.5,  Subsection 5.6.
      3.  Take  three or  more _ readings  for  each  standard metal
solution prepared in step 1.  This can be done either sequential-
ly  or  by alternating the standards, however,  calibration stan-
dards  should  be analyzed at random throughout the  analysis to
check  the calibration  stability.    See  Section 2.8.5,  Subsection
5.7.3.
      4.  Record the instrument response in absorbance'units.   If
using a strip chart  recorder,  take the mid point of the noise as
baseline  and measure   the  net   difference between  the baseline

-------
                                              Section No. 2.8.2
                                              Revision No. 0
                                              Date December 30, 1981
                                              Page 3 of 4

 noise and the peak height.   If the instrument has a digital read-
 out and printer,  there  is  no need to manually measure  peaks  and
 absorbance units can be  read directly.
       5,   Determine the  instrument's  reproducibility by subtract-
 ing the lowest response from  the  highest response,  then dividing
 by.  average  response,  and  multiplying  by 100.   Correct  these
 values by subtracting the  blank.   For  example,  assume  that' the
 three   response  values  derived  from  a 10.0 pg  Pb/ml  standard
 solution  are:

               Absorbance            Peak response  minus  blank
               Blank = 0.005
           1st peak = 0.100              ,      0.095
           2nd peak = 0.098.                   0.093
           3rd peak = 0.098              .      0.093

     The  average  response  value  'for   these  three  analyses  is
0.094,  which  represents  the  standard   metal  solution's concen-
tration  (10.0  Mg  Pb/ml).    The  maximum  percentage absorbance
fluctuation is given by:

      Percent  fluctuation = 10° x  (Q-095 - Q.Q93) _
                                    0.094         ~ z/&-
The  percent  fluctuation  should be <_5%;   if  not,  have the instru- .
ment checked by a qualified service engineer.

-------
                                                     Section No.  2.8.2
                                                     Revision No. 0
                                                     Date  December 30,
                                                     Page  4 of  4
                                                          1981
           'TABLE 2.1.   ACTIVITY  MATRIX  FOR CALIBRATION OF EQUIPMENT
Equipment
Acceptance limits
Frequency and method
   of measurements
                                                                Action if
                                                                requirements
                                                                are not met-
Elapsed-ti'me
 meter
±2 min/24 h
Check against stan-
dard timepiece (known
accuracy) on receipt
and at 6-mo intervals
Adjust, or
replace time
indicator
Timer
±15 min/24 h
Check at purchase
and quarterly against
elapsed-time meter
Adjust, and
repeat test
Orifice cali-
 bration unit
Manufacturer's f1ow
rate = actual  flow rate
±4%
Check flow rate
against primary stan-
dard at receipt and
at 1-year intervals
Adopt new
calibration
curve if no
evidence of
orifice dam-
age; replace
orifice unit
if damaged
Sampler
Q  - Q
 0 „  c x 100 < ±7%
                 Q  = observed flow

                 Q  = flow rate from
                  c   calibration curve
Check against a
calibration orifice
unit on receipt and
after major mainte-
nance of sampler
Rerun points
outside the
limits until
acceptance
limits-are
attained
Atomic absorp-
 tion sepctro-
 photometer
Reproducibility within
±5%
Measure the instru-
ment's, reproducibility
(Subsec 2.2); at pur-
chase check instrument
specifications, align
the  lamp and clean
the  lamp and the
photometer window
 Have  the
 manufacturer
 or  represen-
 tative  ser-
 vice  and ad-
 just  to in-
 strument
 specifica-
 tions

-------
                                             Section No. 2.8.3
                                             Revision No. 0
                                             Date December 30, 1981.
                                             Page 1 of 1
3.0  FILTER SELECTION AND PREPARATION
     Filter selection and preparation must be made carefully for
a  good  sample collection.   The  spectroguality grade  filter,
which  is  commercially available and  contains  the least organic
binders and  inorganic contaminants,   is  recommended,  especially
when  additional  chemical  analyses are  anticipated.    For  more
information,  see Section 2.2.3 of this Handbook.

-------

-------
                                             Section No. 2.8.4
                                             Revision No. 0
                                             Date December 30, 1981
                                             Page 1 of 1
4.0  SAMPLING PROCEDURE
     All criteria for the collection of total suspended particu-
lates2 are equally applicable to the sampling procedure .for lead
(Section 2.2.4 of this Handbook).

-------

-------
                                             Section No. 2.8.5
                                             Revision No.  0
                                             Date December 30,  1981
                                             Page 1 of 14  .
5.0  ANALYSIS OF SAMPLES
     Table  5.1  at the end of  this  section summarizes the major
quality assurance activities for sample analyses.
5.1  Sample Documentation and  Inspection
     1.   Upon  receipt of the  sample  from the field,  remove the
filter folder  from  its shipping envelope and examine the hi-vol
field data  form  (Figure  3.2,  Section 2.213 of this .Handbook.) to
determine  whether  all data  needed  to  verify  the-  sample  for
analysis  and  to  calculate  concentrations  have  been provided.
Void the sample if data are not only missing but also unobtaina-
ble upon inquiry to the field operator or if sampler malfunction
is evident  (e.g., obvious faceplate gasket leakage).
     2.  Record the  filter  number on the hi-vol field data form
and  on the laboratory,  data  log  (Figure' 3.1,  Section. 2.2.3).
     3.  Examine the  shipping  envelope for sample material that
may have become  dislodged  from the filter.  If such material is
observed,  recover  as much as  possible by brushing  it from the
envelope  to the deposit  on  the  filter  with a  soft  camel hair
brush.
     4.  Examine  the  filter for  insects  embedded in  the sample
deposit;  if any are found,  remove them with Teflon-tipped tweez-
ers without disturbing any more  of the  sample  deposit than is
necessary.   If more  than 10  insects are  observed,   refer  the
sample to  the  supervisor for  a  determination whether to accept
or reject it.
     5.  Record  (under remarks)  any  observations  that may have
an impact on the validity of the sample.
5.2  General Analysis Description
     In atomic  absorption  spectrophotometry, the  element being
measured is aspirated into a flame or injected into a carbon arc
furnace and atomized.   A light  beam is  directed through  the

-------
                                             Section No. 2.8.5
                                             Revision No. 0
                                             Date December 30, 1981
                                             Page 2 of 14

flame,  into  a monochromator, and  onto  a detector that measures
the amount of light absorbed by the atomized element.
     Because  each metallic  element has  its  own characteristic
absorption wavelength,  a source lamp  composed of that specific
element is used to minimize spectral or radiation interferences.
The  amount of  absorption of  the  characteristic wavelength is
proportional to the  concentration  of the element in the sample.
     With  the  EPA method, at  least two  types  of interferences
are possible:   chemical and light scattering.3'4'5'6'7  Reports
on the absence of chemical interference far outnumber reports on
its presence; therefore, no correction for chemical interference
is  given  here.   Non-atomic  absorption as  light  scattering,
produced by high  concentrations  of dissolved solids in the sam-
ple, can  produce  a significant  interference,  especially at low
lead  concentrations.    The  interference •is  greater  at  the  '
217.0-nm  wavelength  than at the 283.3-nm wavelength;  in fact,
Scott,  D.R.,  et  al.  have  reported  that  no  interference  was
observed using the 283.3-nm wavelength.3
     In this type  of  photometric analysis,  the concentration of
the  sample and particularly  the  concentration of  reagent  and
standard solutions, are  of utmost  importance to the accuracy of
the determination.  Samples and standard metal solutions must be
carefully and accurately prepared.
5-3  Apparatus
5.3.1   Atomic Absorption Spectrophotometer  - An  atomic  absorp-
tion  spectrophotometer  is  required  for determination  of  lead
content in suspended particulate  matter, and  must  be equipped
with a lead hollow cathode lamp or electrodeless discharge lamp.
5.3.2  Acetylene - The grade recommended by the instrument manu-
facturer should be used as a fuel.   Change  cylinder when pres-
sure drops below 50 - 100 psig.
5.3.3  Air  -  Filtered clearn air (as free  of particulates,  oil
and water as possible) is needed as an oxidant.

-------
                                             Section No. 2.8.5
                                             Revision No. 0
                                             Date December 30, 1981
                                             Page 3 of 14
5.3.4  Glassware -
     a.   Beaker - Beakers,  Borosilicate  glass,  including 30 ml
and 150  ml  are needed  to  digest the  sample.   Phillips beakers
are useful for this purpose.
     b.   Volumetric flask - 100-ml volumetric  flasks  (Class A)
are required for analysis.
     c.   Pipette -  Several  volumetric pipettes  (Class A),  in-
cluding  1,  2,  4,  8,  15,  30, 50 ml should be  available for the
analysis.
     All glassware should be thoroughly cleaned with laboratory
detergent,  rinsed,  soaked  for 4  h  in  20% (w/w)  HN03,  rinsed 3
times  with  distilled deionized water,   and dried  in  a  dust free
manner.
5.3.5   Hot  plate  -  Hot plates,  750  watts,  120 volt,  having
enough plate  surface area for several  sample  beakers,  and cap-
able of  heating to  370°C  (698°F)  are  required  for  the hot ex-
traction of the sample.
5.3.6   Ultrasonication  Bath  Unheated  - An ultrasonication bath
is required for the sample extraction when an ultrasonic extrac-
tion procedure  is employed,  and should  provide  the  'necessary
energy  (>20,000  cycles per second).   Commercially,  available
baths  of 450 watts  or higher  cleaning  power  have been found
satisfactory.
5:3.7  Template - Templates  are  needed to aid in sectioning the
glass  fiber filter (Figure 5.1).
5.3.8  Pizza Cutter  - A pizza  cutter having a thin wheel  (<1 mm
thickness)  is needed to cut the filter.  (Figure 5.2).
5.3.9  Watch glass - A watch glass is needed to cover the beaker
containing the sample.
5.3.10   Polyethylene Bottle  - Linear  polyethylene  bottles  are
needed to store the samples for subsequent analysis.

-------
                                                 Section No.  2.8.5
                                                 Revision  No.  0
                                                 Date December 30,  1981
                                                 Page 4 of 14
                                      30 cm
MANILA FOLDER - TO  PREVENT
FILTER FROM STICKING  TO
PLASTIC
GLASS FIBER FILTER  FOLDED
(LENGTHWISE)  IN  HALF
           12.7
WIDTH OF GROOVE/
1 cm
         ALL  GROOVES
         2 mm DEEP
                                                         IGID PLASTIC
0.8 cm
                                                          JZZA CUTTER
                 25 mm
                (1 in.) WIDE
                                       IDTH OF GROOVE 8 mm
         Figure  5.1.  Sample preparation, filter sectioning 1.

-------
                                       Section No. 2.8.5
                                       Revision No.  0
                                       Date  December 30, 1981
                                       Page  5  of 14
  STRIP FOR
  OTHER ANALYS
                            nim ' 203  T""
                            . 0 A.NALV; ,',
Figure 5.2.  Sample  preparation,  filter sectioning 2.

-------
                                             Section No. 2.8.5
                                             Revision No; 0
                                             Date December 30, 1981
                                             Page 6 of 14              Jfth

5.3.11  Parafilm M  Sealing Film  -  A pliable, selfsealing, mois-
ture  proof,  thermoplastic sheet material,  substantially color-
less  is  recommended for  use in  sealing, the  acidified sample
beakers.   Commercially  available  Parafilm  M  satisfies  this
requirement.
5.4  Reagents (Analysis)      .            •   .
5.4.1   Nitric Acid (HNO3)  Concentrated  -  A.C.S.  reagent 'grade
HNO3 and commercially available redistilled HNO3 have been' found
to have sufficiently low lead concentration:
5.4.2   Hydrochloric Acid (HCl) Concentrated  -  A.C.S.  reagent
grade.        '
5.4.3   Water  -  The same source  or  batch of distilled deionized
water must be used  for all purposes in the analysis.
5.4.4..  3M HNOa  - This  solution is  used in the  hot extraction
procedure-.  To prepare,  add  192  ml of concentrated HNO3 to dis-
tilled  deionized  water  in a  1-2 volumetric flask.  Shake'well,
cool, and dilute to volume with distilled deionized water.  Cau-
tion:   Nitric  acid  fumes are  toxic.   Prepare  in  a well venti-
lated fume hood.
5.4.5.  Glass Fiber Filter  -  Low lead content  of  the filter is
desirable.  EPA typically obtains filters with a lead content of
«75 p.g/filter.  Minimal variation in lead content from filter to
filter is also important.
5.4.6   0.45M  HNOa  - This  solution  is  used as  the  matrix  for
calibration standards when using the  hot extraction procedure.
To prepare,  add 29  ml of concentrated HNO3  to  distilled deion-
ized  water in  a  1-2 volumetric  flask.   Shake well,  cool,  and
dilute to volume with distilled deionized water.
5.4.7   2.6M HNOa +  (0 to 0.9M HCl)  - This  solution  is  used in
the ultrasonic extraction procedure and the concentration of HCl
can be  varied  from  0 to 0.9M.  Directions  for  preparing a 2. 6M

-------
                                             Section No. 2.8.5
                                             Revision No. 0
                                             Date December 30, 1981
                                            - Page 7 of  14

HNO3 + 0.9M HCl solution are as follows: place 167 ml of concen-
trated HNO3 into a 1-2 volumetric flask and add 77 ml of concen-
trated HCl.  Stir  4  to 6 h, dilute to nearly 1 e with,  distilled
deionized  water,  'cool  to  JLOOIII  temperature,  and  dilute  to 1
liter.
5.4.8  0.40M HNOr< + xM HC1  - This solution is used as the matrix
for calibration  standards wnen using  the ultrasonic extraction
procedure.  To prepare, add 26 ml of concentrated HNO:;.. plus  the
ml  of  HCl required  (Equation  5-1),  to a  l-i!  volumetric flask.
Dilute to nearly  1  8. with distilled deionized  water, cool to
room temperature,.and dilute to 1 £.   The amount of HCl required
can be determined from the  following equation:

               y • <77 mlj;<°M151 *
where.                               .
          y = ml of concentrated HCl required, -               /
          x = molarity of HCl from Subsection"5.4.7, and      "
       0.15 = dilution factor from Subsection 5.5.2.
5.4.9  Lead  Nitrate  Pb(NQ:t).^  - A.C.S.  reagent grade  purity of
99.0%.  Heat for 6 h at 120°C and cool in a desiccator.
5.4.10   Stock  Lead Solution (100Q pg  Pb/ml)  in HNO:t  -  Dissolve
1.598 g  of Pb(NO3)2  in 0.45M HNO3 contained in a 1-J> volumetric
flask and dilute to volume with 0.45M  HNO3.
5,4.11  Stock Lead Solution. (1000 yq Pb/ml) in (HNOr,./HCl) -  Pre-
pa,.e as  in 5.4.10 except use the HNO3/HC1 solution from 5.4.84
     Store standard  in  a linear  polyethylene  bottle.   Commer-
cially available certified  lead standard solutions may be used.
This stock solution  may  be  stored up to 2 years.  Label clearly
with contents,  concentration, person  who prepared the  standard,
date prepared and expiration date.  This date should be periodi-
cally checked and a fresh standard made as required.

-------
                                              Section No.  2.8.5
                                              Revision No.  0
                                              Date  December 30,  1981
                                              Page  8  of 14

 5.5    Sample  Preparation for  Atomic  Absorption  Spectrophotometry
 5.5.1  Hot  Extraction Procedure  -
     1.   Gut a 1.9  cm  x 20.3  cm  (3/4  in.  x 8 in.)  strip from
 the  exposed filter using a template and  a pizza cutter 'as de-
 scribed in  Figures  5.1 and  5.2.  Other cutting  procedures  may be
 used.   Care  should be taken  to avoid cross-contamination from
 one  filter  to another by wiping off any fibers which may  adhere
 to  template  or  pizza cutter between samples.   Note:  Lead in
 ambient particulate matter  collected on glass  fiber filters has
 been shown  to be  uniformly distributed across the  filter.3'5'8
 Another  study9  has  shown  that when  sampling near a roadway,
 strip  position contributes• significantly  to the overall  varia-
 bility associated  with lead analysis.   Therefore,, when sampling
 near a roadway,  additional  strips should be  analyzed to minimize
 this-variability.            .
     2.   Fold  the sample  in  half  twice and place. in a  150-ml
 beaker.  Add  15 ml  of 3M HNO3  to  completely cover the sample.
 Cover  the  beaker with a watch  glass.   It is important to keep
 the  sample  covered so that corrosion products  (formed on fume
 hood surface  which may  contain  lead)  are  not  deposited in the
 extract.
     3.   Gently boil the  sample  in  a  beaker on  a  hot plate
 under  a fume  hood  for 30 min.   Do riot  let the sample evaporate
 to dryness.  Caution;  Nitric  acid fumes are toxic.
     4.   After 30 min, remove the beaker from  the hot plate and
cool to near room temperature.  Rinse watch glass and sides of
beaker with distilled  deionized water.
     5.   Decant extract and rinsings into  a 100-ml volumetric
 flask,   and  add distilled  deionized  water  to 40  ml mark  on
beaker, cover with watch glass,  and set aside  for a minimum of
30 min.  This  is a critical step and cannot be omitted since it
allows  the  HNO3  trapped  in  the  filter to diffuse into the rinse
water.

-------
                                              Section No.  2.8.5
                                              Revision No.  0
                                              Date December 30,  1981
   .  '                                         Page 9 -of 14

      6.    Decant the water  from  the filter  into  the volumetric
 flask and  rinse  filter  and beaker  twice  with  distilled deionized
 water and  add  rinsings  to volumetric flask  until  total volume  is
 80 to 85 ml.
      7.    Stopper  flask and  shake "vigorously, and  set aside  for
 approximately  5  min  or  until  foam  has dissipated.
      8.    Bring  solution  to  volume .with distilled  deionized
 water and  mix  thoroughly.  Allow solution  to  settle for one hour
 before  proceeding  with analysis.   (Note:   Do   not  filter  the
 extracted  sample to  remove.particulate matter because of-loss  of
 lead  due to  filtration.  The  final extract  can be  centrifuged  at
 2000  RPM for 30 min  to  remove any  suspended solids.
      9.    If  sample is  to  be  stored for  subsequent  analysis,
 transfer to  a linear polyethylene'bottle,  being carefu.'   not  to
 disturb the settled  solids.
 5-5.2  Ultrasonic Extraction Procedure -             .       .
      1.    Cut  a 1.9  cm x 20.3 cm (3/4 in. * 8 in.)  stup,   as de-
 scribed in Subsection 5.5.1,  step  1.
      2.    Fold  the   sample in half  twice  and place  in a  30-mi
 beaker.  Add  15 ml  of  HNO;J/HC1  solution (see Subsection  5.4.7)
 to  completely  cover the  sample,   and  cover  the  beaker with
 Parafilm.  The Parafilm should be placed  over  the beaker such
 that  none of the Parafilm is in contact with water  in  the'ultra-
 sonic  bath.   Otherwise,  rinsing of  the  Parafilm  in step  4 may
 contaminate the sample.        .              " .
      3.   Place  the  beaker   in  the ultrasonication  bath and
 operate for 30 min.
     4.   Rinse  Parafilm and sides  of  beaker   with  distilled
 deionized water.
      5.   Decant the extract and rinsings into a 100-nrl volumet-
 ric  flask.   Add 20  ml  distilled  deionized water  to covet the
 filter strip,  cover  with. Parafilm, and  set asi-de  for a miinuuiui
of 30 min.   This  is  a  critical  step  and cannot, be omit Led.  The
sample  is  then  processed as   in Subsection  b . s . 1,  yteps  {<.• )

-------
                                             Section No. 2.8.5
                                             Revision No. 0
                                             Date December 30, 1981
                                             Page 10 of 14

through  (9).   Samples prepared by  hot extraction procedure are
now in  0.45M  HN03  and samples prepared by ultrasonic extraction
procedure are now in 0.40M HNO3 + xM HC1.
5.6  Instrument Operation and Analysis
     Because  of the  differences between makes  and models  of
atomic  absorption spectrophotometers,  it is  difficult.to formu-
late detailed instructions applicable* to every instrument.  Con-
sequently, it is recommended that the* user follow manufacturer's
operating instructions.
     1.   Set  the  atomic  absorption spectrophotometer  for  the
standard  conditions  as  follows:  choose  the  correct  hollow
cathode lamp  or  electrodeless  discharge lamp for lead,  install,
and align in the instrument; position the monochromator at 217.0-
nm or 283.3 nm;  select the proper monochromator slit width; set
the light source current according  to the manufacturer's recom- '
mendation;  light the  flame  and  regulate  the  flow of  fuel  and
oxidant; adjust the burner for-maximum absorption and stability;
and balance the photometer.
     2.   If using a  chart recorder, set the chart speed at 8 cm
to  15  cm per minute  and turn on  the power,  servo,  and chart
drive switches.  Adjust the chart pen  to  the  5% division line.
Also adjust instrument  span  using highest calibration standard.
While aspirating the standard sample, span instrument to desired
response.
     3.   The  sample  can  be analyzed  directly  from the volu-
metric flask,  or an appropriate amount of sample decanted into a
sample analysis  tube.   In either  case, care should be taken not
to  disturb  the  settled  solids.   At  least  the  minimum sample
volume  required  by  the  instrument  should  be available  for each
aspiration.
     4.   Aspirate samples,  standards,  and blank into the flame
and record  the  absorbance.   If  using  a recorder wait  for re-
sponse to stabilize before recording absorbance.

-------
                                             Section No. 2.8.5
                                             Revision No. 0
                                             Date December 30, 1981
                                             Page 11 of 14

     5.   Determine the  average  absorbance  value for each known
concentration, and  correct- all absorbance values by subtracting
the blank absorbance value.  Determine the lead concentration in
pg Pb/ml  from the calibration curve as presented in the follow-
ing  subsection.   Record  these values  on the Data  Record Form
(Figure 6.1 of Section 2.8.6).
     Note:
     a.   Samples  that exceed  the calibration  range  should be
diluted with  acid of  the same concentration  and  matrix as the
calibration standards and reanalyzed.
     b.   Check for drift of 'the zero point resulting from pos-
sible  nebulizer  clogging, especially when  dealing with samples
of low absorbance.
5.7  Preparation of Calibration Curve
5.7.1   -Working Standard  Solution  (20 yg  Pb/ml)   -  Prepare .by
diluting  2 ,ml of stock lead .solution .(Subsection' 5.4.10 if the
hot extraction was used  or  Subsection  5.4.11  if the ultrasonic
extraction procedure was  used) to 100 ml with acid  of the same
concentration  and matrix  as  used  in  the  stock  lead  solution.
•This standard should be prepared daily.
5.7.2   A Series  of Calibration  Standards -   Prepare  daily  by-
diluting  the  working  standard  solution  (Subsection  5.7.1)  'as
indicated below with the same acid matrix as used in the working
solution.   Other  lead  concentrations may be  used  provided they
are in the linear range of the instrument.
      Volume of 20 \iq/ml        Final        Concentration
     working standard,  ml     volume, ml        pg Pb/ml
               0    •               100             0.0
              1-0     .            200             0.1
              2.0                 200             0.2
              2.0                 100             0.4
              4.0                 100             0.8
              8.0        .         100             1.6
             15.0                 100             3.0
             30.0                 100             6.0
             50.0                 100            10.0
            100.0         '        100            20.0

-------
                                             Section No. 2.8.5
                                             Revision No. 0
                                             Date December 30, 1981
                                             Page 12 of 14

5.7.3  Calibration Curve  -  Since the working range of analysis,
will vary  depending  on which wavelength Is used and the type of
instrument,  no one  set  o'f instructions  for  preparation  of a
calibration curve can be given.
     Select at least three standards (plus the reagent blank) to
cover  the  linear range  indicated by  the  instrument's manufac-
turer.  Aspirate  these standards and  the  blank  and measure the
absorbance.  Repeat until good agreement is obtained between re-
plicates.  Plot  absorbance  (y-axis) versus  concentration in pg
Pb/ml  (x-axis)  as shown  in Figure  5.3.   Draw  a  straight line
through the  linear  portion of  the  curve, and do  not force the
calibration curve through zero.  Note:   To  determine stability
of the calibration curve, analyze a control standard before the
first sample, after  every subsequent 10th sample,  and after the
last sample.  Vary the control  standard concentration by alter-
nating", in run sequence, a value less  than 1 |jg Pb/ml,  and a  ,
value between 1 and 10 M9 Pb/ml.  If either standard deviates by
more than  5%  from the value predicted by the calibration curve,
take  corrective  action  and  repeat the  previous  10  analyses.

-------
                                               Section No. 2.8.5
                                               Revision No.  0
                                               Date  December 30, 1981
                                               Page  13 of 14
0-20
                      6    3     10    12    14


                       Concentration, ug Pb/ml
16'
18
      Figure 5.3.  Example of a calibration curve for absorbance
                  versus concentration of Pb standard.

-------
                                                                                         1
                                                       Section No.  2.8.5
                                                       Revision No.  0
                                                       Date December 30,
                                                       Page 14 of  14
                                                           1981
               TABLE 5.1.   ACTIVITY MATRIX FOR ANALYSIS  OF  SAMPLES
Activity
Verify documen-
 tation and-
 inspect sample
Atomic •absorp-
 tion spectro-
 photometer
Reayents
Glassware
FiIter strip
Hot extraction
Ultrasonic
 fj/traction
V:»nplfj acid
 concentration
  I if/ration
  .ur /<•
Acceptance limits
Complete documentation;
absence of evidence of
malfunction or sample
loss; ten or fewer
insects visible in
sample
Equipped with lead
hollow cathode lamp or
electrodeless dis-
charge lamp
Al1  reagents must be
A.C.S.  reagent grade
Borosilicate glass and
Class A
Size = 1.9 cm x 20.3 cm
(3/4 in.  x 8 in.)
Do not evaporate to
dryness and cover so
that, corrosion products
are not deposited in
the extract
Provide '20,,000 cycles
per second
0.45M HNO-.; or 0.40M
HNO- + xM HC1
ReprnducifoiIity is
Frequency and method
   of measurement
Visual check
Upon receipt check
for specifications
or certification
Prepare fresh as
introduced in Sub-
sec 5.4
Upon receipt check
for stock number,
cracks, breaks,  and
manufacturer flaws
Check its size
Frequently and
visually check the
level of evapora-
tion
Action if
requirements
are not met
Void sample
Upon receipt check
the label and per-
formance by comparing
to the hot extraction
procedure
Prepare fresh
Recalibrate and
repeat the analysis
Service by
manufacturer
Use new
reagents
Replace or
return to
supplier
Prepare new
strip
Void sample
Return to
supplier or
use a hot
extraction
procedure
Void sample
Check instru-
ment or pre-
pare a new
ca I ibration
curve

-------
                                              Section No.  2.8.6
                                              Revision No.  0
                                              Date  December 30,  IvSI
                                              Page  1  of 6  .
6.0  CALCULATIONS AND DATA REPORTING
     A matrix summarizing the quality control activities  for  the
calculations and the data-reporting requirements is presented in
Table 6.1.   •
6.1  Sample Air Volume
     At  standard temperature  and  pressure (STP)  for samplers
equipped with rotameters:

                    (Qi + Qf)
                V = 	^	 t          -          Equation  6-1

where
           V = air volume sampled, m3,
          9-i = initial air flow rate, m3/min at STP,
          Qf = final air flow rate, m3/min at STP,  and
           t = sampling period (elapsed time),  min.
For samplers equipped with flow recorders:
          V = Qt                                    Equation  6-2
where     Q = average sampling rate, m3/min at STP.
Estimate the Q from the recorder chart.   If the flow rate varies
less than 0.11 m3/min  during the sampling period,  read the flow
rate from the chart  at 2-h intervals and take the average value
for Q.
     Calculation  for  sample air  volume  is  identical  to   the
hi-vol method'(Section 2.2.6 of the Handbook).
6.2  Lead Concentration
6.2.1  Estimation of Lead Concentration of the Blank Filter,   \.tq -
     1.    For testing of the large batches of filters ("500 fil-
ters) select at random 20 to 30  filters  from a given batch.   For
small batches  (<500  filters) a  lesser number of  filters  may bt>
taken.   Cut  one  1.9 cm  -  20.3  cm (3/4  in. x 8  in.)  strip from

-------
                                             Section No. 2.8.6
                                             Revision No. 0
                                             Date December 30, 1981    ^^
                                             Page 2 of 6              {•»

each  filter,  anywhere in the filter..  Analyze all strips, sepa-
rately,  according to the directions  in  Subsections  5.5 and 5.6
of Section 2.8.5.
     2.   Calculate total lead in each filter as
          F  = ua Pb/ml x 10° ml x 12 strips
          Fb   pg Pb/ml x strip  x  filter
where
              F.  = Amount  of lead per  465  square cm (72 square
                   in. ) of blank filter, (jg,
        pg Pb/ml = Lead concentration determined  from Subsection
               5.6 of  Section 2.8.5,
    100 ml/strip = Total sample volume,
12 strips/filter =
       Useable filter  area, 20 cm x 23 cm (8 in.  x 9 in.) _  .
   Exposed area of one strip, 1.9 cm x 20 cm (3/4 in. x 8 in.)
     3.   Calculate the mean,  F^,  and .the relative standard de-
viation (100 x standard deviation/mean).
               n
               Z

                                                    Equation 6-3
where                              .
          F^ = average  amount of  lead  per. 72  square  inches of
               filter, |jg
         F^ • =. amount  of  lead  per  72   square  inches  for each
           1   filter, Mg
           n = number of blank filters analyzed.
The standard  deviation (SD) of the  analyses  for  the blank fil-
ters is given by Equation 6-4,
          SD =
                     n-l
                              1/2
Equation 6-4
The relative standard deviation (RSD) is the ratio

-------
                                              Section No.  2.8.6
                                              Revision No.  0
                                              Date December 30,  1981
                                              Page 3 of 6
           RSD =
 If the relative standard deviation is  high enough so  that in  the
 analyst's  opinion subtraction of  ?b may  result in a  significant
 error in the  jjg- Pb/m3, the  batch should be rejected.   For  ac-
 ceptable batches,  use the value of Fb to correct all lead anal-
 yses  (Subsection 6.2.2)  collected using that batch  of  filters.
 If Fb is below the  lower detectable limit (LDL), no correction
 is necessary.
 6.2.2 Calculation of Lead Concentration of the Exposed Filter  -
 Lead  concentration in the  air sample can  be calculated from data
 tabulated on data record form (Figure  6.1) as follows:
               Pb/ml .x 100 ml/strip x 12 strips/filter) - F
                         '•      7  -              :
where
               C = Concentration, pg Pb/m3,
        Mg Pb/ml = Lead  concentration determined from Section
                   2.8.5, Subsection 5'. 6,
    100 ml/strip = Total sample volume,
12 strips/filter =
  _ Useable filter area, 20 cm x 23 cm  (8 in. x 9 jn. ) _
  Exposed area of one strip, 1.9 cm x 20 cm (3/4 in. x 8 in.)
              Fb = Average  lead  concentration  of blank filters,.
               V = Air volume from Subsection 6.1.
6-2.3   Sample  Calculation of Lead Concentration in Air Sample  -
Data are tabulated on a data record form as shown in Figure 6.1.
The  standard  data  are  recorded  and  a standard  curve is drawn
(Figure 6.2).   The line of best fit is drawn through the points.
Average  all  standards  analyzed  throughout the run but  do  not
include the standards used as checks  of the calibration stabil-
ity.  These are check samples to be compared to the calibration
curve.

-------

                                                   Section  No.  2.8.6
                                                   Revision No.  0
                                                   Date December 30,  1981
                                                   Page 4 of 6
r-oject
Simple location
                                      Date
                                      Ancilyst
Sample
number
/lfi£






Air volume
at STP, m3
A*1I






Avg
blank
V
US'
-^7jr






Absorbance
0. 0
-------
                                            Section No. 2.8.6
                                            Revision No. 0
                                            Date  December 30,  1981
                                            Page  5  of 6
0.20
                          8     10    12    14     16    18   20
                          ug Pb/ml
                Figure 6.2.  Calibration  curve.

-------
                                                    Section  No. 2.8.6
                                                    Revision No. 0
                                                    Date December 30, 1981
                                                    Page 6 of 6
      To calculate total Pb  of a  sample use equation  from Subsec-

 tion 6.2.2.
         _ F(5.7 ug/ml)(100 ml/strip) (12 strip/filter)]  - (0)
                                (2291  m )
      C  = 3.0  (jg Pb/m .
        TABLE 6.1.  ACTIVITY MATRIX  FOR CALCULATION AND  DATA REPORTING
Activity
Calculations
 (1) sample
 volume
 (2) Fb,  blank

 (3) SO and RSD
 of F.  values
     D
 (4) sample
 concentration
Analysis data
 form
Documentation
 and sample
 verification
Documentation
 of report
 data
Acceptance limits
Al1 needed, data
available;  relative
standard deviation of
F.  is not high
All data and calcu-
lations are given
Documentation complete
for calculation of
concentration; all
sample and data
identification numbers
match; absence of evi-
dence of malfunction
or sample loss
All needed data
available
Frequency and method
   of measurements
Visual  check  for each
sample;  repeat all
calculations.
Visual check
Visual check for
each sample
Visual check for
each sample
Action if
requirements
are not met
Void sample;
indicate
errors and
make correc-
tions
Complete
missing data
values
Void sample
Void sample

-------
                                              Section No. 2.8.7
                                              Revision No. 0
                                              Date December 30, 1981
                                              Page 1 of 3
 7.0  MAINTENANCE
      Scheduled or preventive  maintenance  of the sampling equip-
 ment  and  atomic absorption  spectrophotomete_r will result  in a
 reduction-  of downtime  and  remedial  maintenance' requirements.
 Table  7.1  at the  end  of  this  section  summarizes the  quality
 assurance aspects of major maintenance checks.  Record all main-
 tenance activities in a maintenance  log  book.  Normally,  two to
 three  remedial  maintenance  activities are  required per year.
      The maintenance methods  for a.sampler motor,  faceplace gas-
 ket,  rotameter,  sampling head,  motor  gaskets,  and -flow  trans-
 ducer and recorder,  are presented  in  Section 2.2.7-of Volume II
 of  this Handbook.
 7-1 '-Atomic Absorption Spectrophotometer
      As previously indicated, major maintenance and  calibration
 should be done by service  engineers or  qualified operators.   The
 following  general maintenance  procedures  should be carried  out
 only  after  consulting  the manufacturer's manual.
 7-1-1  Light Source  - When problems  are  concerned with a light
 source,  check the hollow cathode  lamp-or electrodeless  discharge
 lamp  mounting bracket,  lamp  connection,  and  make  sure the  in-
 strument is  plugged  in,  turned -on,   and  warmed  up.   If  line
 voltages are low, operate the  power  supply from a variac which
•is  set to  give  maximum voltage.  Lamp  current meter fluctuation
 can  be  reduced  by using  a constant  voltage sine  wave  trans-
 former.

 7'1-2   No Absorbance Response  - Make  sure  that  the  lamp  is
 lighted,  properly aligned, and that  the  wavelength,  slit,   and
 range  controls  are properly  adjusted.   if the  meter  cannot  be
 zeroed,  (1)  adjust the level  of the burner head to avoid  inter-
 cepting  the  light beam,  and  (2)  clean the  lamp and window,  or
 photometer   cover windows,  with  a   dilute solution  of a  mild

-------
                                             Section No. 2.8.7
                                             Revision No. 0
                                             Date December 30, 1981
                                             Page 2 of 3

detergent and  rinse several times with  distilled water.  Dirty
windows or lenses are a major problem when operating the instru-
ment below 2300 A° (23C nm).
7.1.3  Readout Noisy, Flame On - Check the lamp current setting,
fuel ,and oxidizer  flow  rates,  the leviner  to make sure  it is
draining properly,  the nebulizer for  corrosion  around the tip,
the  adjustment  of the nebulizer 'capillary,  the  burner head  (it
may  need cleaning  with  razor blade),  the  acetylene cylinder
pressure, the air pressure, and the air line filter.
7.1.4  Poor Sensitivity (Within 50% of That Suggested in the
Analytical Method Book) -  Check the sensitivity  obtainable  for
several  other elements to  ascertain that the-low sensitivity is
not  due  to  the lamp used.   Check the  slit width,  wavelength,
range  setting,   the  burner alignment,  the  , adjustment of  the.
nebulizer capillary,  the  fuel/oxidant  flow rate ratio to ascer-
tain that it  is optimized for the element to be analyzed.  Make
sure that  the lamp current is not  above the recommended value,
check the lamp  alignment,  and the concentration of the  standard
solution used.
     All other  maintenance problems  such as cleaning of mirrors
or gratings should be discussed with the manufacturer or service
representative.                          •

-------
                                                       Section No.  2.8.7
                                                       Revision  No. 0
                                                       Date  December 30, 1981
                                                       Page  3 of 3
                   TABLE 7.1.  ACTIVITY MATRIX FOR MAINTENANCE
 Equipment
 Acceptance limits
 Frequency and method
    of measurements
 Action  if
 requirements
 are  not met
 Sampler motor
 400  h  operation  of
 motor;  absence of
 malfunction
 Visually check upon
 receipt and  after
 each 400 h of
 operation
                                                                 Replace motor
                                                                 brushes;
                                                                 other  mainte-
                                                                 nance  as  in-
                                                                 dicated
Faceplace
 gasket
Absence of  leaks at
filter seal
 Visually  check after
 each  sampling period
Replace gas-
ket
Rotameter
Absence of foreign
materials; stable
operations
                                          Visually  check  for
                                          each  sample
                       Clean; re-
                       place if
                       damaged
Motor gaskets
Leak tight fit
Visually check each
400 h of operation
                                                                Replace gas-
                                                                kets
Sampling head
Absence of leaks
Visually check each
200 h of operation
                                                                Replace samp-
                                                                ling head
Atomic absorp-
 tion spectro-
 photometer
Absence of dirt and/
or contamination in
light source systems;
absence of current
fluctuation; manu-
facturer's specifica-
tions on sensitivity
achieved
Clean up and adjust
trouble parts, vi-
sually check energy
meter needle; vi-
sually check gauge,
and adjust light
passing systems,
and clean up lamp
and window and
photometer cover
window;  check for
possible troubled
parts as indicated
in Subsec 7.1.4
Replace dam-
aged parts;
use constant
voltage
transformer;
contact manu-
facturer

-------

-------
                                              Section No. 2.8.8
                                              Revision No. 0
                                              Date December 30, 1981
                                              Page 1 of 12
 8.0  AUDITING PROCEDURE
      An  audit  is an  independent assessment  of the  quality  of
 data.  Independence is  achieved by having the audit  made  by  an
 operator other  than  the' one conducting  the  routine  field mea-
 surements and by using audit standards  and equipment different
 from  those  routinely  used.   Routine  quality assurance  checks
 conducted by the operator are  necessary for  obtaining and re-
 porting good'quality data, but  they  are  not to  be  considered  as
 part of the  auditing procedure.
      Based  on the results of  the Reference  Method  Test10 for
 lead  analysis  and Hi-Vol  Sampling  Method2,  three performance
 audits and a system audit  are  recommended and are described  in
 detail in the subsequent sections.
      The  basic  purpose  of  an auditing  program is to  ensure.the
 integrity of the  data  and to assess  the data  in  terms of ac-
 curacy.   Techniques  for estimating the accuracy  of the  data are
 given in  Section  2.0.8 of Volume  II of  this Handbook.  '
 8.1   Performance  Audits
      Performance   audits  are  independent  checks  made  by the
 supervisor or  auditor to evaluate  the  quality of data  produced
 by  the total .measurement  system (sample  location,  sample ana-
 lysis  and data  processing).  Performance audits are normally a
 quantitative appraisal of quality.
     Three performance audits of individual variables  are recom-
 mended:                                           '
      1.  Audit of flow rate calibration
      2.  Audit of lead analysis
      3.  Audit of data processing.
 8-1-1   Audit of  Flow Rate  Calibration  -  The   frequency  of the
 audit of the  flow rate depends  on the  use  of the data  (e.g., for
PSD  air monitoring or for  SLAMS).   it is  recommended that the
flow rate of  each hi-vol sampler be  audited each quarter.

-------
                                             Section No. 2.8.8
                                             Revision No. 0
                                             Date December 30, 1981
                                             Page 2 of 12

     1.   Conduct  the flow  rate  audit  using  a  reference  flow
(ReF) device, a or a similar device.
     2.   Audit the flow  rate at one flow rate.  The ReF device
used for  auditing must be different from the  one used to cali-
brate the flow of the hi-vol sampler being audited.
     3.   Operate  the hi-vol  sampler  at its  normal  flow  rate
with the audit device in place.
     4.   Great care  -must be  used in auditing  the  hi-vol  sam-
plers having flow regulators because "the introduction of resist-
ance plates in the audit device can cause abnormal flow patterns
at the  point of flow sensing.  For  this reason, the orifice of
the  flow  audit device should be used with  a normal glass fiber
filter  in place and  without  resistance  plates, in auditing flow
regulated  hi-vol  samplers,  or other  steps should  be  taken to
assure that flow patterns are not disturbed  at the point of flow
sensin'g.                                 ,
               *   '           •
     5 .   Use  the  known  audit . flow measurement and  the  flow
measured  by the  sampler's normal  flow indicator  to calculate
percent  difference   (Equation  8-1),  a  measure  of inaccuracy.
Both flows  must be referenced to  same temperature and pressure.
Let X. represent the  known flow rate, Y. the measured flow rate,
                                       th
and d. the percent difference  for  the i — audit:
               Y. - X,
          d. =  — =7= - =•  100.                    '     Equation  8-1
          i      X±

Thus if Y. = 52 ft3/min and X. = 50  ft3/min,
     >       •flC!**'              ^
then
         di =
If d-  is greater than ±7%  for  any  one  check,  recalibrate  before
resuming the sampling.
aUSEPA  uses  ReF device with five  orifice plates  that mount onto
 the  faceplate of  the  hi-vol  adaptor;  this  device may be pur-
 chased from Dexco;  Co.,  Inc.,  630 Chapel Hill  Blvd.,  Burling-
 ton, N. C.  27215.

-------
                                              Section No.  2.8.8
                                              Revision No.  0
                                              Date December 30,  1981
                                              Page 3 of 12

      6.    Report the Y^ the X^  and  the .^ on an X-and-R.chart
 Crigure  8.1) under "Measurement Result,  Items  1 and 2."   Record.
 the  d^  in  the  cells preceded  by  the  "Range R."  The d-  can  be
 positive or  negative,   but the  range  is   always  positive;  so
 t/etain the sign -of  the difference since it may  Indicate  trends
 ctnd/or consistent biases.   The steps  in the  construction  of  a
 Duality  control chart, and  the  interpretation of the results are
 in Appendix H,  Volume  I of  this Handbook.11
      7.    Repeat the above  for  each flow rate calibration  audit;
 plot  all points on the  chart;  and connect  the  points by.drawing
 a straight-line.   Tentative limits are ±4.7% (warning lines) and
 x?%   (out-of-control  lines).    Out-of-control   points  indicate
 possible problems in  calibration  errors  or instrument• damage.
 Recalibrate the sampler  prior  to  further  sampling when  out  of
 control.-   After  15  to  20  points  are plotted,  new  control and
 warning  limits  may  be  derived,  as described  in Appendix  H  of
 Volume I of  this  Handbook.11   Do  not increase  the  control and
 warning   limits,  however,  more  stringent   limits  may   be
 established.
 8-L-2  Audit  of Lead Analysis - Each calendar quarter, audit the
 leaa  analysis using glass fiber filters  containing  a, known quan-
 tity  of  lead.  Audit samples are  prepared  by depositing  a lead
 solution  Pb (N03)2,  on 1.9  cm  x  20 cm (3/4 in?  by  8 in.) unex-
 posed  glass fiber  filter  strips, and allowing to  dry thoroughly.
 •It ij  required  that the  audit samples  be prepared using reagents
.-iitferent  from  those, used to generate  the lead  calibration curve
 .in Section  2.8.5.   If  the  routine  network  operators are used  to
 perform  the  audit,  these,  operators  must   not  know  the   audit
 values prior to  the audit. This  means  another  individual must
 administer  the  audit program.   Prepare blind  audit samples  in
 the following concentration ranges:

-------
Section No.  2.8.8

Revision No.  0
Date December 30, 1981

Page 4  of 12
                      rcj
                      O

                      HI


                      S-


                      i
                      i-
                      o
                      M-
                      o

                      4->
                      !=
                      O
                      u
                       3
                      o-
                      co

                      d)
                      «_
                      a
                      en
                          0

-------
                                              Section No.  2.8.8
                                              Revision No.  0
                                              Date  December 30,  1981
                                              Page  5  'of 12

          Range     Cone,  ng Pb/strip    Cone,  pg Pb/m3*
            1            100 to  300       0.5  to 1.5
            2            600 to  1000       3.*0  to 5.0

*Calculation of lead concentration  in ug/m3 is based on sampling
at  1.7  m3/min for  24  h on  an 20  cm  x  25 cm  (8  in. x 10  in.)
glass fiber filter.
     Analyze at least one  audit  sample in  each of  the two  ranges
each  day that  samples  are  analyzed.    If samples are analyzed
only  once per  quarter,  analyze  at least two  audit samples in
each of the two' ranges.  The percentage  difference d between the
audit  concentration  (|jg   Pb/strip)  and  the  analyst's measured
concentration  (ug Pb/strip)  is used to  calculate  analysis inac-
curacy,  (Equation 8-2),
      -     d = cPb(M) - cPb(A) x 100     :                ion 8_2
                  •  UPb(A)
where
           d = Percentage  difference,
      CPb(M) = Concentrati°n measured by the lab analyst,
               (jg Pb/ml, and
      CPb(A) = Audited or  known concentration of audit  sample,
               (jg Pb/ml.
     Tabulate  the  percentage  differences  on  an  X  and R chart
(optional).   The upper and lower control lines and the  upper and
lower warning lines should also be plotted (Figure 8.2) and  used
as  guidelines  to determine  when  results  are  questionable  and
corrective action needs  to be  taken.   Record on  the chart the
nature of the  corrective  action.   Deta-ils for construction  of a
quality control  chart are given in Appendix  H of  Volume I of
this Handbook.ll
     The  recommended  control  limits  for  the  two audit  sample
ranges (0.5 to 1.5  and  3^  to 5 jjg Pb/m;! )  are the 90th percent!le
values for d based  on  the results  of  seven  audits  (8/77,. 1.. 78,
6/78,  1/79,  7/79,  1/80,  and 7/80) performed by the Environmental

-------
Section No. 2.3.8
Revision No. 0
Date December 30, 1981
Page 6  of 12
                          •

-------
                                           .  Section No. 2.8.8
                                             Revision No.. 0
                                             Date December 30, 1981
                                             Page 7 of 12

Monitoring  Systems  Laboratory,  USEPA,  Research  Triangle  Park,
North Carolina.12'13'14'15  By definition, 90% of the laboratory
participants  in  the  audit  obtained values  of d less  than the
values tabulated  below.   The control limits  are  expected  to be
exceeded by 10%  of the  laboratories  to  be  audited,  based on
these seven audits  over  four years.   The 90th percentile values
and  the  known  audit  concentrations  are  given  below  for  each
audit concentration range.
     Audit date
        8/77
        1/78
        6/78
        6/78
        1/79
        7/79
        .1/80
        7/80
0.5 to 1.5 |jg Pb/m3
Known audit
concentration, 90th
|jg. Pb/m3
1.8
0.6
0.4
1.5
1.5
1.2
0.9
0.6
3 to 5 ug Pb/m3
Known audit
concentration, 90th
pg Pb/m3
4.0
5.0
3.5
3.5
4.5
4.2
percentile
°/
/o
45
35
31
. 15
15
16
16
11
percentile
%
23
12
- • 14
13
20
9
for d,



.3
.1

.1
.7
for d,
.5
.5
.8 '
.9
.0
.8
     Audit date
        8/77
        6/78-
        1/79
        7/79
        1/80
        7/80
Based on the results of these seven audits, the recommended 90th
percentile control  limits  for .audit  samples  are  ±16%  for both
the 0.5 to  1.5  |jg  Pb/m3  and the 3.0  to  5.0 pg Pb/m:l concentra-
tion ranges.   The  control limits  of  ±16%  are  also recommended

-------
                                              Section No.  2.8.8
                                              Revision No.  0
                                              Date December 30,  1981
                                              Page 8 of 12

 for  the lower Pb concentrations  range  (0.5 to 1.5- ug Pb/m3) on
 the  assumption that the  first  three audits are not representa-
 tive of current limits  (last  four audits).
     The  method user  should take  part in the EPA semi-annual
 audit  program for lead  analysis.   For more  information on the
 EPA  lead audit  program,  see  Section 2.0.10  of this Handbook.
 8.1.3   Audit  of Data Processing - A  data processing audit  allows
 for  correction of errors, after the  original calculations have
 been  performed.   The  audit rate  of seven measurements out of
 every  100  is recommended.  The audit is  made starting with the
 raw  data on  the  data  form.   When  the original and  the audit
 calculations  do not  agree,  all calculations  for the correspond-
 ing  audit  period  should  be recalculated.   The nature  of  the
 error(s)  should be  clearly explained  to  the  appropriate per-
 sonnel  in order to  minimize their  reoccurrence.   Audit  values
 are recorded  in  the  data log and reported  to the supervisor for
 review.  These  results can be  used  to check computer programs
 and manual methods of data processing.
 8.2  System Audit
     A  system audit  is an on-site  inspection and  review of the
 quality  assurance  system used  for  the  total measurement  system
 (sample  collection,  sample  analysis,  data  processing,   etc.).
Whereas performance audits are a quantitative assessment,  system
 audits are normally a-qualitative appraisal.
     A system audit should be conducted for a monitoring system.
The  auditor should have  sufficient  experience  with  the method
and an  extensive background  with  the characterization technique
 that he  is  auditing.  Figure  8.3  can be  used  as  a preliminary
 form  for use  in  a  system  audit.   These questions  should  be
checked for the applicability to the particular local,  State,  or
Federal agency.  One should also refer to Section 2.0.11 of this
volume  of  the Handbook for  further details  on a  system  audit.

-------
                                                     Section No.  2.8.8
                                                     Revision No.  0
                                                     Date  December 30,  1981
                                                     Page  9 of  12
 I.   What  types of hi-vol samplers are utilized in the network?
                         flloMr   Itinrfc           mock!  (4
 2.   How often are the samplers run? Ka)) daily, (b) once every six days,
     (c) once every 12 days, (d) other — _   . _
 3.   What  type and quality of filter and  number of filters is being utilized?
                   fihts ,   Spec.'h'hftLL&lihj   o
 4.   Are  there any pre-exposure checks  for pin holes or imperfections  run  on
     the  filters?
 5.   What  is the collection efficiency for your filters?
 6.   What  is the calibration procedure for  the hi-vol sampler?
                    i   Procedure  flu-WiruH  \n  5?CJVnn 3.3
 7.   Which  statement most closely estimates  the frequency of flow rate  cali-
     bration?  (a) once when purchased- (b)  once when purchased, then after
     every  sampler modification, or/c^once when purchased, then at regular
     intervals thereafter
 8.   Are  flow  rates measured before and after  sampling period?  Yes y
No
 9.   If the  answer to number 8 is yes using  the equation below, what is  the
     estimated average percent of change  in  the flow rates?

              100 (Q. - Qf)
              	s	 = percent change
                   4i

    ((a)3ess than 10%, (b) 10-20%,  or (c) greater than 20%    .	
10.   Is  there  a  log book at each sampler to  record flows and times?
     Yes      ix^       No
11.   Is  the  atomic absorption spectrophotometer properly calibrated?   U€S
     If  so,  when?   COilihf CL-Hnn  tU/l/e  r\LT\  W}±h  tU-SfU  Samle
12.   Are all  components of the atomic absorption  spectrophotometer correctly
     aligned?  Yes      \^      No _
13.   Qualifications of atomic absorption spectrophotometer operator?   /
14.   Extraction procedure.  (1)  Hot extraction
     (2)  Ultrasonic extraction     •

-------
                                                    Section  No. 2.8.8
                                                    Revision No. 0
                                                    Date December 30,  1981
                                                    Page 10  of 12
15.   Calibration curve check?             OK
16.   Calculation procedure check?  	QK
17.   Are  reagents, calibration  standards, samples,  etc., labeled clearly with
     test numbers, dates,  and all pertinent data?
     Comments:
                Figure 8.3.   Checklist for use by auditor.

-------
                                             Section No. 2.8.8
                                             Revision No.  0
                                             Date December 30, 1981
                                             Page 11 of 12


     The system  audit  should be performed  at the  beginning  of

the monitoring program  and annually  thereafter  unless problems

occur to require  more frequent system audits.

8.3  Activity Matrix

     Table 8.1 summarizes  the quality  assurance activities  for

auditing procedures.

-------
                                                       Section  No. 2.8.8
                                                       Revision No.  0
                                                       Date December 30,
                                                       Page 12  of 12
                                                            1981
                TABLE  8.1.  ACTIVITY MATRIX FOR AUDITING PROCEDURE
 Audit
 Flow rate
  audit
Audit of analy-
 sis process
 using audit
 samples
Data
 processing
 audit
System
 audit
 Acceptance  limits
 di

 Yi
                 Xi
         100
= routinely measured
  flow rate, and

= audited flow rate
d should be within ±16%
for both the 0.5 to 1.5
ug Pb/m3 and the 3 to
5 ug Pb/m3 concentration
ranges (Subsec 8.1.2)
The reported value
should agree with the
audited value within
round-off error
Method described in
this.section of
Handbook
Frequency and method
   of measurement
 Audit each sampler
 quarterly; same
 method as  for cali-
 bration procedure
 Analyze  an  audit
 sample  in each  of
 the two  concen-
 ranges  at
 least once  each ana-
 lysis day and at
 least twice per
 calendar quarter
 that samples are
 analyzed (Subsec
 8.1.2)
 1  in  14 samples or .
 1/mo, whichever is
 greater;  independent
 calculations from
 raw data  to final
 recorded  data
At  the beginning of
a new monitoring
system, and period-
ical ly as appro-
priate; observation
of  procedures and
use  of a check 1ist
Action  if
requirements
are not met
Corrective
action before
resuming sam-
pling; action
noted on X-
and- R chart
Calibration
curve
checked, if
necessary; a
new reference
sample
checked, and
if accepta-
ble, analysis
'resumed; data
accuracy cal-
culated per
Sec 2.0.8
Calculations
for all sam-
ples col-
lected since
previous
audit checked
and corrected
Improved
methods and/
or trainiiH)
programs
ini tiated

-------
                                              Section No.  2.8.9
                                              Revision No.  0
                                              Date December 30,  1981
                                              Page 1 of  1
9.0  ASSESSMENT OF MONITORING DATA FOR PRECISION AND ACCURACY
9.1  Precision
     For  each  monitoring network, collocate  an additional sam-
pler  at  a  minimum  of  one  site  (two  sites  are  required for
SLAMS).

     1.   Select a site with the highest expected geometric mean
concentrations.
     2.   Locate  the  two  hi-vol  samplers  within  4  m  of each
other, but at least 2 m apart to preclude air flow interference.
     3.   Identify one  of the  two  samplers at  the  time of in-
stallation  as  the   sampler   for  normal  routine  monitoring;
identify the other as the duplicate sampler.
     4.   Be sure  that the calibration,  sampling,  and analysis
are  the  same  for the  collocated sampler as  for all  other sam-
plers in the network.
     5.    Operate collocated sampler  whenever  the  routine sam-
pler is operated.
     6.   Use the  differences  in - the  concentrations  (ug Pb/m:: )
between the.routine and duplicate samplers to calculate the pre-
cision as 'described in Section 2.0.8 of this Handbook.
9.2  Accuracy -

     The accuracy of  the  hi-vol method for measurement of Pb is
assessed by auditing a portion  of  the measurement process,  as
described  in  Section  2.8.8.    The  calculation  procedure  for
single instrument accuracy is given 'in Sections 2:8.8 and 2.0.8
of th i .<•; volume  of the Handbook.

-------

-------
                                               Section No. .2.8.10
                                               Revision No.  0
                                               Date  December 30,  1981
                                               Page  1  of 1
  10.0  RECOMMENDED STANDARDS FOR ESTABLISHING TRACEABILITY
      To achieve data of desired quality, two essential considera-
  tions are necessary:   (1)  the  measurement process  must be in a
  state of  statistical  control at the  time  of the measurement and
  (2)  the  systematic errors, when combined  with the random varia-
  tion (errors  of measurement),  must result in an acceptable level
  of  uncertainty.   As evidence in support of  good quality data it
  is  necessary to perform  quality control  checks  and independent
  audits of the measurement process,  to document these data, and to
  use materials,  instruments,  and measurement  procedures  that can
 be traced to an appropriate standard of reference.
      Repeat measurements of standard reference samples,  (primary,
 secondary and/or working  standards)  aid in establishing a condi-
 tion ^ of  process   control.   The  working  calibration  standards
 should be- traceable  to  standards  of  higher  accuracy,   such  as
 those given below.

      1.    A  linearity  test  should be  performed  on the  atomic
 absorption spectrophotometer employing a series of standard metal
 solutions.  This  should be done at  regular  intervals and  when the
 analyst  suspects  erroneous  readings.  Refer  to  Section  2.8.2,
 Subsection 2.2  for details  on instrument performance checkout.
      2.    A positive displacement  rootsmeter is recommended  for
 calibrating the orifice  used to calibrate  the high volume  sam-
 pler.  .See Section  2.2.2,  Subsection  2.5 for  details.
      3.    The elapsed  time  meter,  checked  against an  accurate
.timepiece3 on a semi-annual  basis,  must  be  within ±2 min-per 24 h
 time period.   See  Section 2.1.2,  Subsection  2.4 for details on
 elapsed time  meter  calibration.
      4.    Obtain  a  Pb  (NO3)2  Standard  Reference  Material  (SRM
 928)  from. National  Bureau of Standards.16  Analyze this  standard
 at  regular intervals .along  with samples,  and record accuracy as
 described  in  Section 2.8.8,  Subsection 8.1.2.
 Atomic clock, Boulder, Colorado,  (303) 499-7111.

-------

-------

-------

-------
3EPA
                     Sept. 1982
                              United States
                              Environmental Protection
                              Agency
                              Section 2.9.0

                              Environmental Monitoring Systems
                              Laboratory
                              Research Triangle Park NC 27711
                              Research and Development
                              EPA-600/4-77-027a
Test  Method
                              Section  2.9
                              Equivalent  Method for the
                              Determination  of  Sulfur
                              Dioxide  in the Atmosphere
                              (Fluorescence)
                                                      Outline
                             Section
                             Summary
                             Method Highlights
                             Method Description
                               1.  Procurement of Apparatus
                                  and Supplies
                               2.  Calibration of Equipment
                               3.  Operation and Procedure
                               4.  Data Reduction, Validation, and
                                  Reporting
                               5.  Maintenance
                               6.  Audit Procedure
                               7.  Assessment of Monitoring Data
                                  for Precision and Accuracy
                               8.  Recommended Standards for
                                  .Establishing Traceability
                               9.  Equivalent Method
                              10.  References
                              11.  Data Forms
                            Summary
                              The Reference Method for the
                            determination of sulfur dioxide in the
                            atmosphere (i.e., pararosaniline
                            method) is discussed in Section 2.1 of
                            this Handbook. Many organizations,
                            however, will find it advantageous to
                            conduct continuous monitoring of SO2
                            in ambient air by employing automated
                            monitoring techniques. For use in air
                            quality surveillance systems, state and
                            local agencies are required' to use
                            analyzers that are EPA designated
                            reference or equivalent methods.
                               Documentation
                                    2.9.0
                                    2.9.0
                                    2.9.1
                                    2.9.2
                                    2.9.3

                                    2.9.4
                                    2.9.5
                                    2.9.6

                                    2.9.7

                                    2.9.8
                                    2.9.9-
                                   2.9.10
                                   2.9.11
Number
of Pages
   1.
   1
   5
  10
   7

   1
   2
   1
   1
   1
   1
   4
                              A current list of all designated
                            reference and equivalent methods is
                            maintained by EPA and updated
                            whenever a new method is
                            designated. This list may be obtained
                            from any EPA Regional Office or from
                            the Environmental Monitoring
                            Systems Laboratory, Department E,
                            MD-77, Research Triangle Park, North
                            Carolina 27711. Moreover, any
                            analyzer offered for sale as a reference
                            or equivalent method after April 1 6,
                            1 976, must bear a label or sticker '
                            indicating that the analyzer has been

-------
                         Section 2.9.1
                                      Sept. 1982
 removed from the sample gas stream
 by an appropriate scrubber upstream
 of the reaction chamber. The
 scrubbers may operate at ambient or
 elevated temperature. Certain
 elevated-temperature scrubbers,
 however, have the potential for
 converting ambient hydrogen sulfide
'(which normally does not interfere
 with the fluorescent technique) into
 SOj. In these cases, the hydrocarbon
 scrubber must be preceded by a
 scrubber for H2S.
1.1.2  Specific Fluorescent
Analyzers • Currently, four (4)
instrument manufacturers have EPA
designated equivalent fluorescent SOz
analyzers. The manufacturers and
their respective designated equivalent
methods are: Thermo Electron
Corporation (TECO), EQSA-0276-009;3
Beckman. EQSA-0678-029;4 Monitor
Labs. Inc., EQSA-0779-039;S and
Mcloy (Columbia Scientific Industries
Corp.) EQSA-0580-046.'
  The TECO Series 43 fluorescent SO2
monitor utilizes a pulsed UV light to
excite the S03 molecules.  TECO states
six (6) major reasons for using a
pulsed UV lamp:3
1.  Long bulb life
Z  High intensity—improved signal to
    noise ratio
3.  Small bulb size
4,  Low power requirements—less
    than 1 watt
5.  Long-term stability
6,  Chopped signal processing—no
    dark current  drift
  Before passing into the reaction
chamber the sample air passes
through a permeation dryer, to remove
water vapor, and an aromatic
hydrocarbon cutter (replace every 18
months of operation). The  instrument
operates with a sample  flow rate
between 472 and 1888  cmVmin.
  The Beckman Model 953 fluorescent
SOz monitor uses a continuous UV
light source (deuterium  lamp) but
mechanically chops the  light signal
before it enters the reaction chamber.
  The sample air passes through a
selective scrubber, for the  removal of
rtiS and mercaptans (change every 12
months), and a heated temperature
controlled reactor which removes
pofynuclear aromatic compounds
{replace every 6 months). The sample
then passes into a heated,
temperature-controlled fluorescence
reaction chamber. The chamber is
heated to reduce  condensation of
water vapor. The  instrument operates
vinth a sample flow rate  of 400 to 700
cmVmin.
   The Monitor Labs Model 8850 uses
 UV light from an arc tube to excite the
 SO2 molecules. The UV light passes
 through a mechanical chopp.er before
 entering the reaction chamber. Sample
 air passes through a five (5)-micron
 teflon particulate filter and a catalyst
 (replaced every 12 months) for removal
 of aromatic hydrocarbons before
 entering the heated (40°C) reaction
 cell. The instrument operates with a
 sample flow rate of 500 ± 50
 cmVmin.
   The Meloy Model SA700 fluorescent
 SO2 analyzer operates with a
_ continuous wave of UV light from a
 deuterium lamp. The instrument uses
 a UV detector to monitor lamp
 intensity in the reactor cell and
 compensates and adjusts the UV
 source as the source ages and as
 contamination accumulates on optical
 surfaces. The instrument uses no
 optical or mechanical  chopper.       .
   The sample air passes through a
 membrane dryer to remove water
 vapor and a hydrocarbon scrubber
 must be replaced as part of scheduled
 annual maintenance. The instrument
 operates with a sample flow rate pf
 200 to 500  cmVmin.

 1.2  Strip Chart  Recorder
   Strip chart recorders are
 commercially available with a wide
 variety of prices and specifications.
 Factors to be considered when
 purchasing  a recorder are:
 1.  Compatibility with the output
     signal of the analyzer
 2.  Chart width (minimum of 15 cm is
     recommended for desired
     accuracy of data reduction)
 3.  Chart speed (>2.5 cm/h)
 4.  Response time
 5.  Precision and reliability
 6.  Flexibility of operating variables
     (speed and range)
 7.  Maintenance requirements.

 1.3  Sampling Lines and
 Manifolds
   Sampling lines and manifolds should
 be Teflon or glass to minimize reaction
 with and degradation of the SOz. The
 residence time within  the sampling
 lines should be minimized to reduce
 the possibility of interaction of the SOz
 sample with interim surfaces. If
 particulate filters are employed, they
 should be of Teflon construction.


 1.4  Calibration Equipment
   The recommended calibration
 procedure requires both a permeation
 tube that is traceable to NBS standards
 in a temperature-controlled
 environment and a diluent airstream
 free of SO2 (<0.001 ppm). A detailed
 discussion of this calibration
 procedure appears in Section 2.9.2.
 Calibration may also be conducted by
 diluting an SOz standard gas with zero
 air.
   The calibration system (purchased or
 built) must meet the guidelines
 outlined in the Federal Register.^
 Calibration systems of the types
 described are commercially available.
 Several manufacturers of continuous
 SOz analyzers either offer compatible
' calibration systems or can inform the
 user on where to purchase such
 systems.  When purchasing a
 calibration system, the following
 factors should be considered:
 1.  The permeation tube must be
     traceable to NBS standard
     reference materials (NBS-SRM).
 2.  The method for measuring air flow
     through the calibrator must be
     accurate within ±2% of the actual
     flow.
 3.  The temperature control module
     must be capable of maintaining
     the permeation tube at a
     predetermined temperature within
     ±0.1 °C (0.2°F). The ability to make
     an independent check of the
     temperature within the
     permeation tube chamber is
     desirable.
 4.  The calibrator must be portable.if
     it is to be.used at more than one
     site.
 5.  Maintenance requirements should
     be minimal.
   Permeation tubes are commercially
 available or may be prepared-in the
 laboratory.2'7 The working permeation
 tube  must be traceable to an NBS-
 SRM. If the permeation tube supplied
 with the calibrator is not certified, or if
 the user prepares his own tubes,  the
 user  must conduct certification tests
 and thus purchase an NBS-SRM. The
 following permeation tubes  are
 available as NBS-SRM's:8



SRM
1625
1626
1627




S02
SO2
SO2



Type '
permeation
permeation
permeation




tube
tube
tube

Tube
length,
cm
10
5
2
Nominal
permeation
rate.
/ug/min at 25°
2.8
1..4
0.56



C




-------
                             Sept. 1982
                                                                             Section 2.9.1
  An acceptable protocol for
  demonstrating the traceability of
  commercial permeation tubes to NBS-
  SRM's is described in Section 2.0.7 of
  this volume of the Handbook.
    The user will need a source of zero
  air that is free of contaminants that
  would cause any detectable response
  with the SO2 analyzer. Zero-air is
  commercially available in cylinders or
  can be generated by the user. Because
  fluorescent SOa analyzers may be
  sensitive to the composition of
  synthetically prepared zero-air, a clean
  air system utilizing ambient air may be
  more desirable to use for zero and
  dilution purposes. If ambient air is not
  used, the zero-air cylinder must
  contain the major constituent gases
  normally found in ambient air,
  especially oxygen which is known to
  quench the fluorescence  response.


  1.5  Spare Parts and
  Expendable Supplies
   In  addition to the basic equipment
 discussed above, it is necessary to
 maintain an inventory of spare parts
 and expendable supplies. The
 manufacturer's manual  specifies
 which parts require periodic
 replacement and the frequency of
 replacement. Based on these
 specifications, the management of the
 monitoring network can determine
 which parts and how many of each
 should be available at all times. A
 generalized list of spare  parts and
 expendable supplies is provided below
 (for specific requirements, refer to the
 manufacturer's manual):
   1.  Paniculate filters
  2.  Selective scrubbers for the
      removal of aromatic   .
      hydrocarbons •
  3.  SampJing lines
  4.  Pump diaphragms
  5.  Drier columns
  6.  Activated charcoal
  7.  Recorder chart paper and ink or
      pens
  8.  Calibration gas
  9.  Record forms
 10.  Spare fitjings.
 schedules in Section 2.1.2 of this
 Handbook.

 1.7  Record Forms
   Recordkeeping is critical for all
 quality assurance programs. Standard
 forms similar to those in this
 Handbook should be developed for
 individual programs. Three questions
 to consider in the development and
 storage of record forms are:
 1.  Does the form serve a necessary
     function?
 2.  Is the documentation complete?
 3.  Will the forms be filed so that they
     can be retrieved easily when
     needed?

 1.8   Audit System
   An independent audit system is a
 necessary part of the quality
 assurance program. Two types of audit
 systems may be used:
 1.  A system using an NBS traceable
    permeation tube (Subsection 1.4),
    or
 2.  A dynamic dilution system with a
    tank of SO2 certified traceable to
    an NBS-SRM or a commercially
    available Certified Reference
    Material (CRM) and a zero-air
    supply (Section 2.5.6).
In either case, the system used for
auditing must not be the same as that
used to calibrate the analyzer.  ,
1.6   Reanalysis of Calibration
Working Standards
  All working standards for
calibrations should be reanalyzed at
least once every 6 mo. (Subsection
7.2.6 of Section 2.0.7 describes the
procedures for analysis and for
reanalysis of permeation devices).
Flow-measuring devices should be
reca-librated using the procedures and

-------
                  Section 2.9.1
                                                         Sept. 1982
          .1
          (3

               Q
               Q
               O
               ^r
       41


       I
               to
                            OQ
1
•§
5
               e
      a. o c
                   Oa

                   o.
          a
          ,3.


          1
          Q

-------
                         Sept. 1982
                                              Section 2.9.1
Table 1.1.    Activity Matrix for Procurement of Equipment and Supplies
Equipment and
supplies
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Fluorescence
SOz ana/yzer
Strip chart
recorder
Samp/ing fines
and manifolds
Calibration
devices
SOz permeation
tube
Zero -air -~
Record forms
Audit system
Performance according
to specifications in
Table 4.1, Sec 2.0.4
Compatible with output
signal of analyzer;
chart width of 1 5 cm
(6 in.) is recommended;
accurate chart speed
Constructed of Teflon
or glass
Must meet guidelines
of Reference 1
Traceable to NBS-SRM;
meets limits in trace-
ability protocol for
accuracy and stability
(Sec 2.O.7;
•Clean dry ambient air.
free of contaminants
that cause detectable
responses with the SOz
analyzer
Standard form
developed
Must not be the same
system as used for
calibration, either an
NBS traceable perme-_ ,
at/on tube or a dynamic
dilution system (Sub-
sec 1.8)
Have the manufacturer
provide a strip chart
recording of specific
analyzer's performance;
verify performance
specifications at
installation
Check upon receipt
As above
See Reference J
Analyze against an
NBS-SRM; protocol
in Sec 2.0.7
.See Sec 2.9.2 .
N/A
Check the system
against a known
standard
Have the
manufacturer
make proper
adjustments;
recheck the
performance
Return to
supplier
As above
As above
Obtain new
working stan-
dard; check
for traceability
. Obtain 'air
from another
• source or re-
generate
Revise forms
as appropriate
Locate problem;
correct, -or
return to
supplier

-------

-------
                             Sept. 1982
                                        Section 2.9.2
                                    2.0   Calibration  of Equipment
   The accuracy and precision of data
  derived from the air monitoring
  equipment are dependent on the
  quality assurance procedures used,
  primarily the dynamic instrument
  calibration. Calibration determines the
  relationship between the observed and
  the true values of the variable being
  measured. Table 2.2 at the end of this
  section summarizes the quality
 • assurance activities for calibration.
   Dynamic calibration involves-
  introducing gas samples of known
  concentrations into an instrument in
  order to adjust the instrument to a
  predetermined sensitivity and to derive
  a calibration relationship. This
- 'relationship is derived from the
  instrument's responses to successive
  samples of different known
  concentrations. Introducing these
  standard gas mixtures in decreasing
  order of concentration will minimize
  the response times. As a minimum,
 three reference points and one zero
 point are recommended to define this
 relationship. Linearity of fluorescent
 analyzers is'-also checked at this time.
 The true value of the calibration gas.
 must be traceable to NBS-SRM's
 (Section 2.0.7).
   Most currently available monitoring
 instrument systems (e.g., the
 fluorescent SOz analyzer) are subject
 to drift and variation in internal
 parameters, and thus cannot be
 expected to maintain accurate
 calibration over long periods of time.
 Therefore,  it is necessary to
 dynamically check the calibration
 relationship on a predetermined
 schedule. Precision is determined by a
 one-point check at least once every 2
 weeks. Accuracy is determined by a
 three-point audit once each quarter.
 Zero and span checks (Subsection'
 3.4.2) must be used to document.
 within-control conditions; these
 checks are  also used in'data reduction
 and validation.
2.1    Calibration Gases
  The recommended method of
dynamically calibrating a fluorescent
SOz analyzer requires both a certified
permeation tube traceable to an NBS-
SRM in a temperature-controlled
environment (±0.1 °C) and diluent air
void of SO2  «0.001 ppm). To conduct
biweekly precision checks and Level 1
zero and span checks, the  user will
need a supply of zero air, a cylinder of
  SO2 (50 to 100 ppm) in nitrogen, and a
  dynamic dilution system or a
  calibration system with a permeation
  tube capable of generating the
  precision check point at 0.08 to 0.1
  ppm and a span check point at 70 to
  90 percent of the analyzer's
  measurement range. To implement a
  quality assurance program for
  calibration, the user will therefore
  need the following:
  1.   An SOz permeation tube or device
      that is traceable to an NBS-SRM
  2.   Zero air
  3.   An SO2 span gas that is traceable
      to an NBS-SRM or commercially
      available CRM
  4.   A calibration system!

  2.1.1  SO2 Permeation Tubes  - The
  NBS-SRM's provide a reference
  against which all calibration  gas
  mixtures must be compared. See
  Section 2.9.1 (Subsection 1.4) for an
  address for obtaining a list of NBS-
  SRM's which are available for S02
  analyzers.
   One function of NBS is to supply
  standards, but they do not supply
  working calibration gases. Therefore
  the user is advised either to purchase
  commercially available certified
  permeation tubes that are traceable to
  NBS standards or to make the tubes.2
  In either case, the user is responsible
 for the verification and reanaiysis of
 working standards versus NBS-SRM's.
 or CRM's. Procedures the user must
 follow to verify working calibration
 gases are outlined in Section 2.0.7. •

 2.1.2 Dilution Air - Zero-air, free of
 contaminants which  could cause a
 detectable response in the fluorescent
 SO2 analyzer, must be used for the
 calibration, the precision check,  and
 the Level  1 zero and span checks. This
 air is used to establish the analyzer
 zero base line and to dilute the SOz to
 the required concentrations. Zero-air
 may be supplied from cylinders or from
 a clean air system.
  Because the fluorescent reaction
 has some degree of sensitivity to
 aromatic hydrocarbons, COz levels.
 and to the oxygen/nitrogen ratio,3 it is
 recommended that a clean air system
 be used. The air for this system must
 be drawn from outside the station to
prevent excess CO2 levels. Water
vapor and aromatic hydrocarbons
should be removed from the zero air.
    If compressed air cylinders are used,
  the air.should have the following
  properties for use with fluorescent
  analyzers:
  1.   The same O2 and N2 percentage
      composition as ambient air
      (20.94% Qz. 78.08% N2).
  2.   A C02 content similar to that of
      ambient air (between 300 and 400
      ppm).
  3.   Less than 0.1 ppm aromatic
      hydrocarbons.
  2.1.3.  SOzPrecision and Span
  Cases - Aluminum or steel cylinders
  containing 50 to 100 ppm S02 in N2
  are available from most specialty gas
  suppliers. Aluminum cylinders have
  been demonstrated by NBS to have
  superior stability for storing SO2
  mixtures, and they are preferred
  whenever possible. These gases can
  be diluted to the desired concentration
  by using zero-air and a dynamic
 dilution system. The cylinder gas
 concentration should be certified to
 NBS-SRM or commercially available
 CRM (Certified Reference Materials)
 using EPA Traceability Protocol No. 2
 (Section 2.0.7). NBS-SRM'or
 commercially available CRM at 50 and
 100 ppm SO2 in N2 should be used for
 the traceability analysis. A CRM may
 be used directly for precision or span
 checks. However, due to the limited
 supply of NBS-SRM, an SRM should
 not be used directly for routine
 precision or span checks. A list of gas
 manufacturers who have approved
 CRM's is available by writing to:
  U.S. Environmental Protection
   Agency
  Environmental Monitoring Systems
   Laboratory (MD-77)
  Research Triangle Park, North
   Carolina 27711
  ATTN: List of CRM  Manufacturers
(Note: CRM's  are cylinder gases
prepared by gas manufacturers
according to an NBS-EPA procedure to
within ±1% of existing SRM
concentrations. Each CRM lot of 10 to
50 cylinders is audited by EPA. Each
lot receives written approval from NBS
and this approval must accompany any
CRM sold.)
  Precision and span gases  may also
be generated  by a calibrator using an
SOz permeation tube traceable to an
NBS-SRM.

-------
                          Section 2.9.2
                                                                             Sept. 1982
 2.2  Calibration System
   The calibration system consists of
 two primary parts.
 1.  The temperature controlled
     permeation device
 2.  A dynamic dilution system.

 2.2.1  Temperature Controlled
 Permeation System • The purpose of a
 permeation system is to generate a
 low SOj concentration at a constant
 rate. This is done by holding a
 permeation tube at a constant
 temperature (±0.1 °C) for which the
 permeation rate is known. The SO2
 permeating from this tube is carried
 away by a low flow of gas (usually
 clean dry air or N2) to a mixing
 chamber where it is  accurately diluted
 with zero air to the concentration
 desired.

 2.2.2  Dynamic Dilution System - A
 dynamic dilution system  is required to
 dilute the SOj output from either the
 temperature-controlled permeation
 system or an 80s gas cylinder to the
 desired concentration. All parts in
 contact with the SO2 output must be
 glass or Teflon. The system must be
 capable of controlling and measuring
 flow rates to within ±2% of stated
 flow.
 2.3   Dynamic Multipoint
 Calibration Principles
   Dynamic calibration involves
 introducing gas samples of known
 concentrations to an instrument in
 order to adjust the instrument to a
 predetermined sensitivity and to derive
 a calibration relationship. A minimum
 of three reference points and one zero
 point uniformly spaced covering 0 to
 80 percent of the operating range are
 recommended to define this
 relationship.
  The recommended method of
 dynamically calibrating an SO2
 analyzer requires a certified
 permeation tube traceable to an NBS-
 SRM in a temperature controlled
 environment (±0.1 °C) and diluent air
 that is free of SO2 «0.001 ppm).
 Temperature must be verified with an
 NBS traceable thermometer prior to
 calibration.
  The permeation tube is  held at a
 constant temperature for a minimum
 ol 24 hours to allow the SO2 to diffuse
 from the tube at a known  rate. The low
 How of zero air that is passed over the
 permeation tube  serves as a carrier for
 the SOa. This purged air is then diluted
wuh different quantities of zero air to
generate the desired concentrations.
   The analyzer's recorded response is
  compared with the known
  concentration to derive the calibration
  relationship. This relationship is used
  to convert the analyzer's responses
  during sampling into ppm's of SO2.
  The recorded response may be either
  voltage output or percent chart (%
  chart) as long as it is consistent with
  that used to determine the calibration
  relationship.
   Biweekly precision checks are used
  to calculate  the variability of the
  calibration relationship over a period of
.  time.  Three-point audits conducted
  quarterly are used to check the
  analyzer's accuracy. These precision
  checks and the 3 (or 4) point audits are
  used to generate precision/accuracy
  data for the  reporting organization.
  They are not intended for use in
  reducing or validating data since they
  are performed infrequently. Level 1
  zero and span checks must be used to
 'document within-control conditions
  and to validate the collected data.

  2.4  Calibration Procedures
   The procedures for multipoint
  calibration of an SO2 analyzer by an
  SO2 permeation system are specified
-  in the.Federal Register.' To facilitate
  these procedures, operational and
  calculation data forms have been
  developed as aids in conducting
  calibrations and quality assurance
  checks.  Detailed descriptions of the
  calibration theory and procedures for
  SO2 permeation systems are in the
 Federal Register.'
   Documentations of all data on the
 station, instrument, calibrator,
 reference standard, and calibration
 procedures are of prime importance
 since the validity of the data collected
 by the instrument is dependent on the
 quality of the calibration. Calibration
 must be performed with a calibrator
 that meets all conditions specified in
 Subsection 2.2.

 2.4.1   General Calibration
 Recommendations - It is important that
 the  fluorescent analyzer be operated
 during calibration under conditions
 identical to those during normal
 ambient air sampling. No modifications
 or alterations shall be made to the
 analyzer's components, flow system,
 prescribed flow rate, or other
 parameters. Concentrations of SO2
 intended for calibration must be
 generated continuously by means
 entirely independent of the analyzer.
 The flow rate of the calibration gas
 must exceed the sample flow rate of
 the analyzer. The calibration gas
 should flow through a manifold, and
 the analyzer should draw its sample
 through the regular ambient air
 sampling line, which is attached to a
 port of the vented calibration manifold.

 2.4.2  Calibration Procedure for SOz •
 The procedure for multipoint
 calibration of a fluorescent analyzer is
 necessarily general. That given here is
 for fluorescent analyzers equipped
 with a linearized output. Where
 analyzer-specific explanations are
 necessary, the reader is referred to the
 manufacturer's instruction manual.
  The following procedure using the
 forms shown in Figures 2.1  and 2.2, is
 given to aid in the collection and
 documentation of calibration data.
  1.   Record the official name and
      address of the station  on the
      form; where appropriate, the
      name and address should be the
      same as that appearing on the
      SAROAD site identification form-
      to eliminate any confusion by
      persons not familiar with the
      station.
  2.   Identify the analyzer being
      calibrated by the manufacturer's
      name, model, and serial number.
  3.   Identify the person performing
      the calibration and the date of
      calibration.
  4.   Identify the calibrator used. If
      the calibrator was purchased,
      record the manufacturer's name, .
      model  and se'rial number.
      Calibrators assembled by the
      user should be assigned an
      identification number so that
      calibrations can be referenced to
      that particular apparatus.
  5.   Identify, by supplier and tube
      number, the reference standard
      to  be used. Provide a record of
      NBS traceability for any tube
      used in a calibration and include
      the data of verification and the
      name of the person who verified
     the reference standard.
  6.   Identify the device used to
      measure the flow of the dilution
     air.
  7.  Record the barometric  pressure
     and the.shelter temperature.
  8.  Record the analyzer sample air
     flow.
  9.  Record the zero and span knob
     settings after the calibration is
     completed. (These settings can
     be  used as a basis of comparison
     when changes are later
     determined in the instrument
     performance.)
10.  Record the temperature at which
     the permeation tube is
     maintained during calibration
     and use the recorded

-------
                                                                            Section 2.9.2
Sept. 1982
span drift exceeds its respective
limit, investigate the cause of the
drift, take corrective action, and
calibrate the analyzer. Individual
agencies may wish to use limits
which are tighter than those  in
Table 9.1.

-------
                          Section 2.9.2
                                                                                Sept. 1982
                                                 Calibration Data Forms
                                                          2.
                                                   5r
                                                                                       "7 ' " O cL
 4,  Calibrator used
 5.  SO) standard _
                                            ur     60 *.
                                                           Ti//ag
    verified against NBS-SRM
    BV        Tom
                                 f (ff 2.W	T\J&£   /C.Q
                                                           Date
  e,
 7,  Barometric pressure
                                                      AvA&LF
                               f&i-
                                                    mm Hg Shelter temperature
 8,  Anafyter sample flow readings

 9,  Zero knob setting _:	
                                                          Span knob setting
 10,  Permeation equilibrium temperature .
                               A
                                             2.5". O
    Permeation rate (PR)
                                                    fjg/min
1},   Temperature at which air flow rate was 'measured (A T)

12.   Vapor pressure of water at temperature (A T)	

                                                       *• '•  '
Equation 2-1
                                 238
                     BP*VP
STP correction factor = - X
                       760    AT T 273
Equation 2-2
Ft » F X STP correction factor
                                                                                                                5Q
                                                                                                          mm Hg
                PR   MV
                - X -
                Ft     M
Squatton 2-3
     .__ ,
ppm ISOiJ out


equation 24
Response (% scale, =  fS°j£ur   X ,00 - Zso,
                                                 Calibration Equations

                                                         BP = barometric pressure, mm Hg
                                                         VP = vapor pressure of water, mm Hg at AT (Table 2. 1)
                                                         AT = temperature at which air flow rate was measured. °C
                                                                                       '
                                                           F = uncorrected flow rate
                                                          F T = total air flow rate, corrected to 25°C'and 760 mm Hg. L/min
                                                         PR = permeation rate at equilibrium temperature, fjg/min
                                                         MV = molecular volume of SO2 at 25°C and 760 mm Hg 24.45 L/molj
                                                                                  Q2 (64 g/mol)
                                                                            _„,.,-. /n
                                                                            — O.382 L/g
                                                           M = molecular weight
                                                          V  _  24.45 L/mol
Figure 2, 1,   Example of a calibration data form.
                                                         ..
                                                         M        64 g/mol
                                                          URL = upper range limit of analyzer
                                                          Zso, - recorder response to zero air

-------
                                Sept.  1382
                                                                    Section 2.3.2
                                                 SO2 Calibration and linearity check.

Calibration
points
Zero
80% URL
1-
2
1
F.
L/min

f.5~
-------
                            Section 2.9.2
                                                                                  Sept. 1982
Calibration
point
Zero
80% URL
1
2
: 3
4
X
O-OOO
O-VOO
0.00,2
o.ov^


X2
o-ooo
OJbQ
O.OY/
O.OO2.


y
• S".O
8S.O
ViT^
M.O


y*
JS
-I**.*
AObl
t*(i


xy
o.o
J+0
9.2
O'b


 x * concentration, ppm      y = recorder reading. % scale

                   ^=0.203    *v=
 7/>e equation of the tine fitted to the data is written as
                                                  Y = y + bfx - x} = (V - bx) = bx = a + bx
 where
       Y — the predicted mean response for the corresponding i.


       b = the slope of the fitted line.


       a « the y-intercept.


       n = the number of calibration points.

                  l£*l (Tvl
            txy -
Figure 2.3.    Calculation form for the method of least squares.

-------
                           Sept. 1982
Section 2.9.2
  1
  O

  5-
  
-------
Section 2.9.2
                              10
Sept. 1982
Table 2. 1,
Temp.
°C
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
3O
31
32
33
34
' 35
Tabla 2.2.
Calibration
activities
Saturation Vapor Pressure Over Water (°C. mm. HgJ
0.0
12.788
13.634
14.530
15.477
16.477
17.535
18.65O
19.827
21.068
22.377
23.756
25.209
26.739
28.349
30.043
31.824
33.695
35.663
37.729
39.898
42.175
0.2
12.953
13.809
14.715
15.673
16.685
17.753
18.880
20.070
21.324
22.648
24.039
25.509
27.055
28.680
3O.392
32.191
34.082
36.068
38. 155
4O.344
42.644.
Activity Matrix for Calibration
0.4
13.121
13.987
14.903
15.871
16.894
17.974
19.113
20.316
21.583
22.922
24.326
25.812
27.374
29.015
30. 745
32.561
34.471
36.477
38.584
4O.796
43.117
Procedures
Acceptance limits
0.6
13.290
14.166
15.092
16.071
17.105
18.197
19.349
20.565
21.845-
' 23.198
24.617
26.117
27.696
29.354
31.102
32.934
34.864
36.891
39.018
, 41.251
43.595

0.8
13.461
14.347
15.284
16.272
17.319
18.422
19.587
20.815
22.110
23.476
' 24.912
26.426
28.021
29.697
31.461
33.312
35.261
37.308
39.457
41.71O
44.078

Action if
Frequency and method requirements
of measurement are not met
Permeation
tube
Dilution gas •
Span gases
Multipoint
calibration
Traceable to NBS stand-
ards
Zero-air free of con-
taminants; Sec 2.0.7,
Subsec 7. 1. and TAD10
Cylinder gases cer-
tified to NBS-SRM or
commercial CRM cylin-
der gas or to an NBS-
SRM permeation tube
Cases generated by
calibrator using an
SOz permeation tube
Calibration proce-
dure in Subsec 2.2.
and the Federal
Register,1 completed
Figs 2.1 and 2.2
Subsec 2.0.7 for
frequency and method
TAD10
Assay against an
NBS-SRM semi-annually;
Sec 2.0.7
Perform at least
once every quarter,
or anytime a level
span check indicates
a discrepancy, or
after maintenance
which may affect the
calibration; Subsec
2.5
Return to
supplier, or
make another
permeation tube
Return to
supplier, or take
corrective action
with generation
system
Working as
standard is
unstable and/ or
measurement
method is out
of control; take
corrective action
(e.g., obtain new
span gases)
Repeat the
calibration

-------
                            Sept 1982
                                       Section 2.9.3
                                    3.0   Operation and Procedure
   Essential to quality assurance are
 scheduled checks for verifying the
 operational status of the monitoring
 system. At least once each week the
 operator should visit the site. Every
 two weeks. Level 1 zero and span
 checks must be made on the analyzer.
 Level 2 zero and span checks should
 be conducted at a frequency desired by
 the user. Level 1 and 2 checks,are
 described in-depth in Section  2.0.9 of
 this Handbook.
   At least once every two weeks, an
 independent precision check at a
 concentration between 0.08 and 0.10
 ppm SC>2 must be conducted. Table 3.1
 at the end  of this section summarizes
 the quality assurance activities for the
 routine operations discussed in the
 following sections.
   For documentation and
 accountability of activities, a checklist
 should be compiled and then filled out
 by the field operator as each activity is
 completed. A simplified example
 checklist.js given in Figure 3.1. A more
 comprehensive check list should be
 developed  for specific sampling
 stations. -                     "   -
   In Subsections 3.1 and 3.2,
 reference is made to the sampling
 shelter and the sample inlet system,
 but the design and construction  of
 these components  are not within the
 scope of this Handbook. For more
 information refer to an  in-depth  study
 of these in Reference 11.
3.1   Shelter
  The shelter's role in quality
assurance is to provide a temperature-
controlled environment in which the
sampling equipment'can operate at its
optimum. The mean shelter
temp.erature should be between 22°
and 28°C (72° and 82°F). A
thermograph should be installed at.the
shelter to continuously record daily
fluctuations in temperature.
Fluctuations greater than ±2°C (±4°F)
may cause the electronic components
of the analyzer to drift and may
introduce error into the data.
Fluctuations outside of these limits
should be  identified, and the data for
the affected time period should be
flagged to  indicate possible
discrepancies.
 3.2  Sample Introduction
 System
   The sample introduction system
 consists of an intake port, the
 paniculate and moisture traps, the
 sampling manifold and blower, and the
 sampling line to the analyzer. The field
 operator, as part of the quality
 assurance program, should inspect
 each of these components for
 breakage, leaks, and buildup of
 paniculate matter or other foreign
 objects; check for moisture deposition
 in the sample line or manifold; see that
 the sample line is connected to the
 manifold; see that any component of
 the sample introduction system that is
 not within tolerance is either cleaned
 or replaced immediately. See Section
 2.0.2 for more details.

 3.3   Recorder
   During each weekly visit to the
 monitoring site, the field operator
 should use the following list to check
 the recorder for proper operation:
 1.  Ink trace for legibility
 2.  Ink level  in reservoir •
 3.  Chart paper for supply
 4.  Chart speed control setting
 5.  Signal input range switch
 6.  Time synchronization.
 Any operational parameter that is not
 within tolerance must be corrected
 immediately.

 3.4   Analyzer
  'Specific instructions in the
 manufacturer's manual should be read
 thoroughly before attempting to
 operate the analyzer. As part of the
 quality assurance program, each  site
 visit should include a visual inspection
 of the external parameters of the
 analyzer; the  zero and span checks;
 and a biweekly precision check when
 applicable.

3.4.1  Visual Inspection - The field
operator should inspect the external
operating parameters of the analyzer;
these will vary from instrument to
instrument, but in general they will
include the following:
 1.   Correct settings of flow meters
    and regulators.
 2.   Cycling of temperature control
    indicators.
 3.   Temperature level, if equipped
    with a pyrometer.
 4.  Verification that the analyzer is in
     the sampling mode rather than the
     zero or the calibration mode.
 5.  Zero and span potentiometers
     locked and set at proper values.

 3.4.2  Zero and Span Checks - Zero
 and span checks must be used to
 document within-control conditions
 and to provide interim checks on the
 response of the instrument to known
 concentrations. A quality control chart
 can be used to  provide a visual check
 to determine if  the analyzer is within
 control conditions. If a response is
 outside of the prescribed limits, the
 analyzer is out  of control and the
 cause must be determined and
 corrected. A zero check should be
 conducted at the same time that the
 span check is performed.
   Level 1 and Level  2 zero and span
 checks are recommended and must be
 conducted i'n accordance with
 Subsection 9.1  of Section 2.0.9. Level
 1 zero and span checks must be
 conducted every two weeks. Level 2
 checks should be conducted between
 the-Level 1 checks at a frequency
 desired by the user.  Span
 concentrations  for either Level 1 or 2
 checks should be between 70% and
 90% of the measurement range. The
 data should be recorded on the zero
 and span check form, Figure 3.2.
  Zero and span checks are used to
 provide:
 1.  Data to allow analyzer adjustment
    for zero and span drift
 2.  A decision  point for calibrating the
    analyzer
 3.  A decision point for invalidating
    the monitoring data.
 Items 1 and 2 are described in detail in
 Subsection 9.1.3 of Section 2.0.9 and
 item 3 is described in Subsection 9.1.4
 of the same section.
.  When the response from a  span
 check is outside of the control limits,
 the cause for the extreme drift should
 be determined,  and corrective action
 should be taken. Some of the causes
 for drift are:
 1.  Lack of preventive maintenance
 2.  Fluctuations in electrical power
    supply
 3.  Major fluctuations in sample flow
 4.  Change in zero air source
 5.  Change in span  gas concentration
 6.  Degradation of photomultiplier
    tube

-------
                         Section 2.9.3
                                                                            Sept. 1982
 7.  Degradation of UV light source
 8,  Electronic and physical
    components not within
    manufacturer's specifications.
 Corrective actions for the above can be
 found in the manufacturer's
 instruction/operations manual.

 3.4.3  Precision Check - For
 continuous analyzers,  periodic checks
 are used to assess the data for
 precision. A one-point precision check
 must be carried out at least once every
 2 weeks on each analyzer at an SO2
 concentration between 0.08 and 0.10
 ppm. The analyzer must be operated in
 its normal sampling mode, and the
 precision test gas must pass through
 all filters, scrubbers, conditioners, and
 other components used during normal
 ambient sampling. The standards from
 which precision check test
 concentrations are obtained must be
 traceable to NBS-SRM's or NBS/EPA-
 approved commercially available
 Certified Reference Material (CRM).
 Direct use of a CRM as a working
 standard is acceptable, but direct use
 of an NBS-SRM as a working standard
 is discouraged because of the limited
 supply and expense of SRM's.
 Standards used for calibration may
also be used. The precision check
procedure is as follows:
 1.  Connect the* analyzer to a
    precision gas that  has a
    concentration between 0.08 and
    0.10 ppm. An SO2 precision gas
    may be generated  by an SO2
    permeation tube or by dilution of a
    high concentration (50 to 100
    ppm) SO2 standard gas. If a
    precision check is  made with a
    zero and span check, it must be
    made prior to any zero or span
    adjustments.
 2,  Allow the analyzer to sample the
    precision gas until a stable trace is
    obtained at the recorder.
 3.   Record this value on the precision
    check data form (Figure 3.3), and
    mark the chart as "unadjusted
    precision check." Information from
    the check procedure is used to
    assess the precision of the
    monitoring data; see Section 2.0.8
    for procedures for calculating and
    reporting precision.

-------
                              Sept. 1982
                                                                                     Section 2.9.3
 Site ID .
          ool
 Site name
             1.  Inspect thermograph for temperature variations greater than ±2°C (4°FJ. Identify time frame of any temperature level
           2.   Inspect sample introduction system for moisture, particulate buildup, foreign objects, breakage, and leaks.

               Comments:. J-C&f/M t/^sfr?/)*/  OP
        .  3.   Check to see if sample line connected to manifold.

               Comments:
           4.   Inspect data recording system.



               •  Legibility of trace

               •  Ink supply

               •  Paper supply

               •  Chart speed selector

               •  Signal input range switch

               •  Time synchronization

               Comments:  	
OK
                                                                   Corrective
                                                                   action taken
5.   Inspect analyzer's operational parameters



    •  Sample flow rate

    •  Oven temperature light flashing

    •  Analyzer in sample mode

    •  Zero and span potentiometers locked at
        correct setting

    Comments:  .
                                                                 OK
           Corrective
           action taken
          6.   Zero the analyzer

          7.   Check to see if unadjusted zero is within tolerance.

               Comments:     ZSgQ   OKAV  A.T  
-------
                            Section 2.9.3                      4                     Sept. 1982
          12,    Record cylinder pressure of zero and span tanks.



                       Zero air    /&*> F5/
         , 13,    Close valves on zero and span tanks.
Signature of technician
Figure 3,1.  •  Example of an operational checklist (backside).

-------
                          Sept. 1382
                                                                        Section 2.9.3
 Site ID
             OOI
Location,


Address
 Adjusted zero.
                                                      Pollutant
                                                                    50;
                                                      Analyzer
                                                    _ Serial number _


                                                    .Adjusted spa/7.
                                                                          V7/O.  O
                                                                                OF
   Date
             Time
                      .Operator
                                     Unadjusted
                                        zero,
                                       % chart
                                                               Span
                                                           concentration,
                                                               ppm
Unadjusted
  span,
 % chart
Z-l-So
                       &4C.
                                                                                      8V
Figure 3.2.   Example of a Level 1 zero and span check data form.

-------
                           Section 2.9.3
                                                             Sept. 1982
 Site ID
001
 Location
           J336»
                                      _ Pollutant _




                                       Analyzer
                          A/g.     Serial number       W/Q. O
Date
2-/-£o
Time
/O30
Operator
4&C.
Precision
test gas
concentration,
ppm
O.IO
-
Analyzer
response.
% chart
H
ppm
o-ol?
Difference",
ppm
-0.005"
'Difference » analyzer response - test gas concentration.




figure 3,3.   Example of precision check form.

-------
                           Sept. 1982
                                                Section 2.9.3
Table 3.1.    Daily A ctivity Matrix
Characteristic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Shelter temper-
ature
Sample intro-
duction system
Recorder
Analyzer oper-
ational set-
tings
Analyzer oper-
ational check
Precision
check
Mean temperature be-
tween 22° and 28°C
(72° and 82°F). daily
fluctuations not
greater than ±2°C (4°Fj
No moisture, foreign
material, leaks, or ob-
structions; sample line
connected to manifold
Adequate ink and chart
paper; legible ink
traces; correct
settings of chart
speed and range
switches; correct
time
Flow and regulator
indicators at proper
settings; temperature
indicators cycling or
at proper levels;
analyzer set in •
sample mode; zero and
span controls locked
Zero and span within
tolerance limits;
Subsec 9. 1.3 of Sec
2.0.9
Precision assessed as
described in Sec 2.0.8
and Subsec 3.4.3'
Edit thermograph
chart daily for
variations greater
than ±2°C (4°F)
Visually inspect
weekly
Visually inspect
weekly
Visually inspect
weekly-
Check level J zero
and span every 2
weeks; check Level 2
between Level 1
checks at frequency
desired by user
Check every 2 weeks.
Subsec 3.4.3
Mark strip chart
for the affected
time period;
repair or adjust
temperature
	 control system
Clean, repair.
or replace
as needed
Replenish ink
and chart paper;
adjust recorder
time to agree
with clock;
note on chart
Adjust or repair
as needed
Isolate source
of error, and
repair; then
recalibrate
the analyzer
Calculate;
report precision;
Sec 2.0.8

-------

-------
                           Sept. 1982
                                                                           Section 2.9.4
                       4.0   Data Reduction, Validation,  and Reporting
   This section is the same as that of
 Section 2.5.4. Table 4.1 summarizes
 the quality assurance activities for the
 data reduction, validation, and
 reporting.
 Table 4.1.    Activity Matrix for Data Reduction, Validation, and Reporting
Activity
Acceptance limits
Frequency and method
of measurement
 Action if
 requirements
 are not met
Data reduction
Data validation

 Span drift
  check
Strip chart
 edit
Data reporting
                           Stepwise procedure,
                           Subsec 4.1
Level 1 span drift
check <25%, Sec
2.0.9
No sign of malfunc-
tion
Data transcribed to
SAROAD hourly data
form; Re f 13
                                  Follow the method
                                  in Subsec 4.1
Check at least every
2 weeks; Sec 2.5.3;
Ref 12 recommends
screening procedures
to identify gross
anomalies
Visually edit each
strip chart; Subsec
4.2
                                                            Visually check
                                  Review the
                                  reduction
                                  procedure
Invalidate data;
take corrective
action; increase
frequency of
Level 1 checks
until data
are acceptable

Void data for
time interval
for which
malfunction
detected

Review the
data transcribing
procedure

-------

-------
                           Sept. 1982                   1                   Section 2.9.5
 5.1   Preventive Maintenance
   Because maintenance requirements
 vary from instrument to instrument,
 the supervisor should refer to the
 manufacturer's manual for a specific
 analyzer. After becoming familiar with
 these requirements, the supervisor
 should develop a suitable preventive
 maintenance schedule.

 5.2   Corrective Maintenance
   Corrective maintenance is defined
 as nonscheduled activities that
 become necessary due  to system
 malfunctions. A few examples of
 corrective maintenance are: replacing
 a damaged pump diaphragm; cleaning
 a clogged sampling line; and replacing
 the selective scrubber for aromatic
 hydrocarbons. The need for corrective
 maintenance becomes apparent as the
 operator performs the operations
 described in Section 2.9.3. When the
 need for corrective maintenance
 arises, the operator should refer to the
 owner's manual for troubleshooting
 procedures. A detailed record of
 corrective maintenance activities
 should be kept on file for each analyzer
 at the site to identify recurring
 malfunctions; maintenance log
 appears in Figure 5.1.
  Catyf/on: When replacing, aligning-,
 and otherwise servicing the deuterium
 source lamp, always wear UV-
 absorbing glasses to protect the eyes
 from the ultraviolet radiation produced.
 Ordinary prescription spectacles with
glass lenses are suitable. Plastic
 lenses may not provide adequate
 protection.
                                           5.0   Maintenance

-------
                    Section 2.9.5
                                                           Sept. 1982
            OOl
Sue number

Sit* location /TWcififtf	

a.***,   333fa JeFre*5o^
                                            - Pollutant
                                            .Instrument
                                                        ACME.   Fluorescence.
                                            . Serial number
                                                           288-70*1^-5
   Date
               Technician
  Event
 initiating
maintenance
Maintenance
  activity
                                                                           Comments
 2-/-80
                             Loss erf-
                            d/iArf def/<2. ef /ort
                          UV
                                                                    •ft.ll
                                                                                 UV
                                                                            source.
Figure 5.1,   Analyzer maintenance log.

-------
                          Sept. 1982
                                                                         Section 2.9.6
                                     6.0  Auditing  Procedure
  Table 6.1 summarizes the quality
assurance activities for audits. This
section is the same as Section 2.5.6.
See References 14 and 15 for the
frequency and brief descriptions of
audit procedures.
Table 6.1.    A ctivity Matrix for A udit Procedure
Audit
Acceptance limits
Frequency of method
of-measurement
Action if
requirements
are not met
Multipoint
calibration
audit
Data reduction
audit
Systems audit
Difference between
measured and audit.
values is used as mea-
sure of accuracy: Sec
2.0.8
Step wise procedures for
data reduction. Subsec
6.2; no audit dif-
ference exceeding
±0.02 ppm
Method in this sec-
tion of the Handbook
Perform at least once
a quarter; Subsec
6.1.3 for procedure
Perform independent
data processing check
on a sample of the
recorded data; check
1 day of every 2
weeks of data, 2
hours each day
At startup of new
monitoring system,
and periodically
observe as appropri-
ate; checklist.
Fig 6. 4
Recalibrate
the analyzer
Check all re-
maining data
if one or
more data re-
duction checks
exceed ±0.02
ppm
Initiate
improved
methods and/
or training •
programs

-------

-------
                          Sept. 1982                   1                   Section 2.9.7
            7.0   Assessment of Monitoring Data for Precision and Accuracy
  For continuous analyzers, in
 SLAMS, NAMS, or PSD networks a
 biweekly check is performed to
 determine if the measurement process
 is within control and to assess the data
 for precision. These data can be used
 to calculate estimates of single
 instrument precision, and reporting
 organization precision as prescribed in
 Section 2.0.8 of this volume of the
 Handbook. The precision check      '       .                                  .
 procedures described in Section 2.9.3,        '                -
 Subsection 3.4.3 are consistent with
 those in References 14 and 15.
  Estimates of single instrument
 accuracy as well as reporting
 organization  accuracy for ambient air
quality measurements from
continuous methods are based on the
results of the in-depth accuracy audit
and are calculated according to the
procedure in  Section  2.0.8. The audit
is described in Section 2.9.6.

-------

-------
                           Sept. 1982                   1                   Section 2.9.8
                8.0   Recommended  Standards for Establishing Traceability

   Tp achieve data of desired quality,
 two considerations are essential:
  1.  The measurement process must
     be in statistical control at the time
     of the measurement, and
 2.  The systematic errors, when
     combined with the random
     variation in the measurement
     process, must result in an
     acceptable uncertainty.
 As evidence in support of good quality
 data, it is necessary to perform quality
 control checks and independent audits       .
 of the measurement process; to
 document these data (e.g., by means of
 specific data forms or a quality control
 chart); and to use materials,
 instruments, and measurement
 procedures that can be traced to
 appropriate standards of reference.
   Data must be routinely obtained by
 repeat measurements of standard
 reference samples (primary, secondary,
 and/or working standards), and a
 condition of process control must be
 established. The working standards
 must be traceable to either NBS-
 SRM's or commercially available
 CRM's, such-as those listed below:

  NBS-SRM's Available for Use in Establishing Traceability of Permeation Tubes8



SRM
1625
1626
1627



Type
SOz permeation tube
SOa permeation tube
SO2 permeation tube

Tube
length,
cm
10
5
2
Nominal
permeation
rate.
fjg/min af25°C
2.8
1.4
0.56
    NBS-SRM's Available for Use in Establishing Traceability of Compressed
                             Cylinder Gases

                                                             Nominal
    bHM _ ; _ Type                     concentration
      o                        S02inN2          "             s
    1694 _ SQ2inN2 _ IQOppm

  A list of gas manufacturers who
have approved CRM is available by
writing to:
  U.S. Environmental Protection
   Agency
  Environmental Monitoring Systems
   Laboratory (MD-77)
  Research Triangle Park, North
   Carolina 2771 1
  ATTN: List of CRM  Manufacturers

-------

-------
                            Sept. 1982                   1                   Section 2.9.9
                                       9.0   Equivalent Method
   A method description is not given
  herein. The concepts of equivalent
  analyzers are discussed in Section
  2.0.4 of this volume.of the Handbook.
  The analyzer must also comply with
  the performance specifications in
  Table 4.1 of  Section 2.0.4. An
  instruction manual including the
  calibration procedure must accompany
  the analyzer when- it is delivered to the
  purchaser. This instruction manual
.  has been reviewed and approved by
  EPA as part of the equivalency test
  program. The user of the analyzer
  should use the method description in
  this section of the Handbook and the
  instruction manual.
   A list of equivalent methods may be
 obtained from any EPA regional office
 or from the EnvironmentalMonitoring
 Systems Laboratory, Department E,
 MD-77, Research Triangle Park, N.C.
 27711. Any analyzer offered for sale
 as an equivalent method after April 16,
 1976, must bear a label indicating this
 designation by EPA.

-------

-------
                            Sept. 1982
                                     Section 2.9.10
   1.   Code of Federal Regulations 40.
       Protection of the Environment.
       Parts 50 to 69. Revised July 1,
       1977.
   2.   Summary of Performance Test
       Results and Comparative Data
       for Designated Equivalent
       Methods for S02, EPA Document
       No. QAD/M-79.12.
   3.   Thermo Electron Corporation,
       Environmental Instruments
       Division. Instruction Manual
       Model 43 Pulsed Fluorescent
       SO2 Analyses Equipped with .an
       Aromatic Hydrocarbon Cutter.
       TE5405-112-77, Revision C.
       Hopkinton, Massachusetts.
   4.   Beckman Instruments, Inc.
       Beckman Model 953 Fluorescent
       Ambient Sulfur Dioxide
    '   Analyzer. Fullerton, CA. May
       1979.
   5.   Monitor Labs, Inc. Monitor Labs,
       Inc. Model 8850 Fluorescent
       SOz Analyzer Instruction
       Manual. Document 8850 Rev. D.
       San Diego, CA. September
       197,9.
   6.   Columbia Scientific Industries
      Corp. Fluorescent Sulfur Dioxide
      Analyzer Model SA700
      Operation, Maintenance, and
      Parts Manual.. Meloy
      Laboratories, Inc. Springfield,
      Virginia. 1980 and 1981.
   7.  Scaringelli, F. P., O'Keefe, A. E.,
      Rosenberg, E. and Bell, J. P.,
      "Preparation of Known
      Concentrations of Gases and
      Vapors with Permeation Devices
      Calibrated Gravimetrically",
      Analytical Chemistry, 42 871
      (1970).
.  8.   Catalog of NBS Standard
      Reference Materials.  NBS
      Special Publication 260. 1981 -
      83  Edition. U.S. Department of
      Commerce, NBS, Washington,
    '  D.C. November 1981.'
  9.   Quality Assurance Handbook for
      Air Pollution Measurement
      Systems. Vol. I. EPA-600/9-76-
      005. March 1976.
 10.   Use of the Flame Photometric
      Detector Method for
      Measurement of Sulfur Dioxide
      in Ambient Air, A Technical
      Assistance Document, EPA-
      600/4-78-024, May  1978.
 11.   Field Operations Guide for
      Automatic Air Monitoring
      Equipment. U.S. Environmental
     10.0   References

     Protection Agency, Office of Arr
     Programs; October 1972.
     Publication No. APTD-0736, PB
     202-249, and PB 204-650.
12.  U.S. Environmental Protection
     Agency, Sreening Procedures for
     Ambient Air Quality Data. EPA-
     450/2-78-037, July 1978.
13.  AEROS Manual Series, Volume
     II: AEROS Users Manual, U.S.
     Environmental Protection
     Agency, Research Triangle Park,
     N.C., EPA-450/2-76-029,
     OAQPS No. 1.2 - 039, December
     1976.
14.  Appendix A - Quality Assurance
     Requirements for State and
     Local Air Monitoring Stations
     (SLAMS); Federal Register. Vol.
     44, No. 92, pp. 27574-27582,
     May 1979.
1 5.  Appendix B - Quality Assurance
     Requirements for Prevention of
     Significant Deterioration (PSD)
     Air Monitoring, Federal Register,
     Vol. 44, No. 92, pp. 27582
     27584, May 1979.

-------

-------

-------

-------
                   Addendum to Section 2.10
Reference Method for the Determination of Participate Matter as PM10
         in the Atmosphere (Dichotomous Sampler Method)
This section is up-to-date. The blank data forms that are mentioned have
been removed.

-------
T,

-------
                                                            Section No.:  2.10.0
                                                            Date:   April 11,  1990
                                                            Page:   1
SECTION 2.10

Reference Method for the Determination of  Particulate Matter as  PM10  in  the  Atrnos-
pnere (Dichotomous Sampler Method)
Subsection

    0
    1

    2
    3
 •   4
    5

    6
    7
    8
    10
    11
    12
           Title

 Introduction
 Procurement of  Equipment  and
   Supplies
 Calibration Procedures
 Field Operations
 Filter  Preparation  and  Analysis
 Calculations, Validations,  and
   Reporting of  PM10  Data
Maintenance
Auditing Procedures
Assessment of'Monitoring  Data for
   Precision and Accuracy
Recommended Standards for
   Establishing Traceability
Reference Method
References
Data Forms
Full Section
Number

  2.10.0

  2.10.1
  2.10.2
  2.10.3
  2.10.4

  2.10.5
  2.10.6
  2.10.7

  2.10.8

  2.10.9
  2.10.10
  2.10.11
  2.10.12
No. of
 pages
    7
   17
   14
   10

    7
    6
   12
    1
    5
    1
  10
                                                        C Printed on Recycled Paper

-------
                                                           Section No.:  2.10.0
                                                           Date:  April  11, 1990
                                                           Page:  2
2.10.0  INTRODUCTION
     As described in 40 CFR 50, Appendix J, the reference method for PM10 sampling
(i.e., sampling particulate matter with an aerodynamic diameter less than or equal
to a nominal 10 /im), a PM10 sampler draws a measured quantity of ambient air at a
constant flow rate through a specially designed particle size discrimination inlet.
Particles in the PM10 size range are then collected on one or more filters during
the specified 24-hour sampling period.  Each sample filter is weighed before and
after sampling to determine the net weight" (mass) gain of the collected PM10
sample.  ,

     The total volume of air sampled is determined from the measured volumetric
flow rate and the sampling time.  The concentration of PM10 in the ambient air is
computed as the total mass of collected particles in the PM10 size range divided by
the volume of air sampled.  This sampled volume must be corrected to EPA reference
conditions (25 *C, 760 mm Hg or 101 kPa), and PM10 data are expressed as micrograms
per standard cubic meter (//g/std. m3).  The particle size discrimination character-
istics (sampling effectiveness) of the sampler inlet over the PM10 size range,  and
particularly the particle size at which the sampling effectiveness is 50%, are
functional specifications tested in accordance with explicit procedures prescribed
in 40 CFR 53.  Sampling methods for PM10 that meet all requirements in both Parts
50 and 53 are designated as PM10 reference methods for use in SLAMS and PSD moni-
toring.  These designated methods are usually identified by the name of the manu-
facturer and by the model of the sampler.

     Two types of samplers that meet designation requirements are the high volume
PM10 sampler (HV PM10) and the dichotomous sampler.  Only the dichotomous sampler
is discussed in this section of the Handbook; the HV PM10 sampler is discussed in
Section 2.11.

     The most common commercially available dichotomous samplers are low flow rate
(16.7-L/rain) samplers that collect particles with an aerodynamic diameter up to a
nominal size of 10 /»m.  (Note:  In reference to PM10 samplers, all particle sizes
are specified by their aerodynamic rather than physical diameter.)  Dichotomous
samplers further divide the sample into fine (0- to 2.5-^m) and coarse (2.5- to
W-ftm) fractions, which are collected on separate filters.

     Particles with aerodynamic diameters greater .than 10 /tm are removed from the
air sample by inertial separation tn a specially designed fractionating in.let such
as the one illustrated in Figure 0.1.  Particle-laden air is drawn into the inlet
and deflected downwards into the acceleration jet of an impactor.  Because of their
greater inertia, particles larger than 10 /
-------
                                                         Section  No.:  < 2.10.0
                                                         Date:  April  11, 1990
                                                         Page:  3
AMBIENT  AIR FLOW
            Vent Jet
                                                        Upper Plenum
     Upper Flange


        Screen


  Lower Flange

Deflective Cone
                                                       Acceleration Jet
                                                      Middle Plenum
                                                     Lower Plenum
                     FLOW TO VIRTUAL IMPACTOR
          Figure 0.1. Example of a PM10 dichotomous sampler Inlet head.

-------
                                                                   Section No,:   2.10.0
                                                                   Date:   April  Jl, 1990
                                                                   Page:   4
                                            From Aerosol Inlet
               FIn« Particles,
               <2.5>im

            Coarse Particles,
Fitter Cassette
      Fine
      Particle
      Fitter
                                                                         inlet Tube
             Virtual
             Impactor
             Nozzle
             Virtual
             Impactor
             Receiver Tube
             Fitter
             Cassette
             Coarse
             Particle
             Fitter
           High Velocity
             Air Row
                                 To Control Module
Low Velocity
  Air Flow
           Figure 0.2. Principle of the secondary (2.5 ^m) particle size separation
                       In a dlchotomous sampler by virtual impactlon.

-------
                                                           Section No.:  2.10.0
                                                           Date:  April 11, 1990
                                                           Page:  5


collected on the coarse particle filter, a correction must be made when fine and
coarse particle concentrations are calculated.

Method Highlights

     The procedures provided in this document are designed to serve as guidelines
for the development of quality assurance (QA) programs associated with the opera-
tion of dichotomous samplers.  Since recordkeeping is a critical part of QA activi-
ties, several data forms are included to aid in the documentation of necessary
data.  The blank data forms (Subsection 12) may be used, as they are,  or they may
serve as guidelines for preparing forms more specific to the needs of the indi-
vidual monitoring agency.   Partially filled-in forms are included at appropriate
places in the text to illustrate their uses.

     Tables at the end of  some sections summarize the material  covered in the text
subsections.  The material  covered in the various subsections of this section is
summarized here:

     1.    Subsection 1,  Procurement  of Equipment and Supplies,  describes recom-
          mended procurement procedures,  equipment selection criteria,  and  minimum
          accuracy requirements.   It also provides an example of a permanent
          procurement record.

     2.    Subsection 2,  Calibration  Procedures,  provides  detailed calibration .
          procedures for the dichotomous  sampler;   References are provided  for
          calibration procedures  for the  flow-rate transfer standards and other
          monitoring equipment.   A table  is provided at  the end of this  subsection
          that summarizes the  acceptance  .limits  for calibration.

     3.    Subsection 3,  Field  Operations,  details  procedures for filter  installa-
          tion,  performance of operational  quality control  (QC)  checks,  sample
          handling,  and  data documentation.   Complete documentation of background
          information during sampling  is  one of  several QA activities that  are
          important  to future  data validation; particularly important are any
          unusual  conditions existing  during collection of the  sample.   Such
          conditions should be noted.

    .4.    Subsection 4>  Filter Preparation  and Analysis,  presents important consid-
          erations for the  handling,  integrity,  equilibration,  and weighing of
          filters.   A high  qua-lity filter  is  recommended  for use  when additional
          chemical analyses  are expected.   Subsection  2.10.4  also briefly describes
          minimum  laboratory QC procedures.   The analytical  balance must be cali-
          brated annually,  and the filters  must  be  equilibrated  in a  controlled
          environment.

     5.    Subsection 5,  Calculations, Validations,  and Reporting  of PMIO Data
          presents calculations for determining  PMIO mass concentrations and mini-
          mum data validation requirements.   The final data  review and validation
          including  standardized reporting procedures, are  all  important parts of'a
          QA program.  Independent checks of  the data and calculations are required
          to ensure  that the reported data are both accurate and  precise

-------
                                                      Section No.:
                                                      Date:  April
                                                      Page:  6
 2.10.0
11,  1990
6.   Subsection 6, Maintenance, recommends periodic maintenance schedules to
     ensure that the equipment is capable of performing as specified.  Minimum
     maintenance requirements and procedures are outlined.  The objective of a
     routine maintenance program is to increase measurement system reliabil-
     ity.,

7.   Subsection 7, Auditing Procedures, presents independent audit activities
     and laboratory evaluations that provide performance checks of flow-rate
     measurements and data processing.  Filter weighing procedures and balance
     operation evaluations and a system audit checklist are also provided.
     Independent audits evaluate data validity.

8.   Subsection 8, Assessment of Monitoring Data for Precision and Accuracy,
     describes the assessment procedures for determining the accuracy and
     precision, of the data.  The precision check is performed by using collo-
     cated samplers.

9.   Subsection 9, Recommended Standards for Establishing Traceability,  dis-
     cusses the traceability of monitoring equipment to establish standards of
     higher accuracy,  a necessary prerequisite for obtaining accurate data.

10.   Subsections 10 and 11 contain the PM10 Reference .Method and pertinent  '
     references, respectively,  used to prepare this document.  Subsection 12
     provides  blank data forms for the convenience of the user.

-------
                                                            Section No.:  2.10.1
                                                            Date:  April 11, 1990
                                                            Page:  1


 2.10.1  PROCUREMENT OF EQUIPMENT AND SUPPLIES

      The establishment of an ambient PM10 air monitoring network requires the pro-
 curement of specialized equipment and supplies for field operations'and subsequent
 filter analysis.  Information in this section has been provided to assist the
 agency in selecting the proper equipment.  Subsection 1.1 presents minimum sampling
 equipment necessary to conduct field operations.  Recommended laboratory instrumen-
 tation is presented in Subsection 1.2.            '

      In addition to field operations and laboratory equipment,  a data handling
 system (including forms,  logs,  files, and reporting procedures)  must  be developed
 and implemented.                                           .                   K

      It is  recommended that each agency establish minimum monitoring  equipment
 requirements  and budgetary limits before the procurement procedures are initiated.
 UE°nireEeipt  ?f th! samPlin9 equipment and supplies,  appropriate procurement  checks
 should be conducted to determine their acceptability,  and whether they are  accepted
 or  rejected should'be  recorded  in a  procurement  log.   Figure  1.1 is an example, of
 such  a log.   This log  will  serve as  a permanent  record for procurements and provide
 fiscal  projections  for future programs.   It  will  also  help to provide the continu-
 ity of equipment and supplies.   Table 1.1,  at  end of  the subsection   lists  the
 major equipment needed, how it  should be tested,  suggested acceptance limits   and
 actions  to be  taken  if acceptance limits are not  met.

 1.1   Procurement Prerequisites—Field Operations   •         '

 1.1.1   Dichotomous  Samplers  -                                         -

      The  individual sampler  must  meet U.S. EPA operational  standards  and be a  model
 designated as  a  reference or equivalent  method.   A complete listing of minimum
 sampler  requirements (i.e.,  40 CFR 50, Appendix J) is  contained  in the reference
 method  reproduced in Subsection  10.   Dichotomous  samplers  not designated as refer-
 mlln?  nJ  fSU1wa]ent 1"eAh^S maV not be used for Deporting  data to determine attain-
 ment  of  the National Ambient Air  Quality  Standard  (NAAQS)  for particulate matter.
 Cost  for dichotomous samplers will vary with the manufacturer and the  sophistica-
 tion  of  the sampler.   Basic  considerations include the flow control and measurement
 system, maintenance requirements, reliability, and ease of operation.

      Using only  one model  of sampler  in a network will minimize the variety of   '
 spare parts  required to keep the  network  in operation.  An in-house inventory  of   '
 general maintenance supplies and  replacement parts is  recommended.  These include
 various -hand tools  general  all-purpose cleaner,  penetrating oil, distilled water
 Kimwipes (or equivalent),  soft brush, and cotton swabs.  Spare parts for the
 sampler may  be supplied by the manufacturer or many may be purchased locally.

 1-1.2   Calibration Equipment -

ha  <«a)1-brati?n activUies require specialized equipment that will not necessarily
be used in routine monitoring.  At a  minimum, the following equipment   is required:

-------
                                                   Section No.:  2.10.1

                                                   Date:  April 11, 1990

                                                   Page:  2
    w


    01



    o
    o


    o
     B
    w
    O
    O
ur
    •a

    "8
    •2
    o
    oc
    o

    i
    UJ
o
S »-

-------
                                                             Section  No.:   2.10.1

                                                             Date:  April  11,  1990

                                                             Page:  3
                            abJe °f a"urate1y measurin9 ambient temperatures to  the

           Technolnn   /N     eferen"d *° a National  Institute of Standards and
           Technology  (N ST  or an American Society for Testing Materials  (ASTM)

           thermometer within ,2'C (NIST is the former National Bureau of  Standards
                               °f acc"rate1y measuring' barometric pressure over a
                          a^Hg (66i?° 106 kPa) t0 the nearest -"ill^e? of Hg
                      r   ? Jeast annual]y to a standard of known accuracy within
           h™     •  °r 1aboratory measurements,  a Fortin-type, mercury-coTumn
           barometer is appropriate.  For field measurements,  a portable  aneroid
           barometer (e.g.,  a climber's or engineer's altimeter)  is appropHate

                  ^ transfe>:istanda'"ds capable of  accurately  measuring the total
                 and coarse  flow rates  of a dichotomous  sampler.   Tables 21  and 2 2

                     aLK

                          ~
                        ^^^


                              ^
          An adapter of the correct dimensions that will connect the transfer
1.1.3  QC Flow-Check Device

                  A, a

-------
Section No.:
Date:  April
Page:  4
 2.10.1
11, 1990
         «
       If
       «~
       21
       « £
       M _

       £*
        .8
       c9 a

       JJ*
       i|>
       MOO
       &3S.

       IP
       « a =

       2?5
       flj ^ «W
      _?  fw.

       «"* ^
       g c tj
      j;: 
-------
                                                            Section No.:  2.10.1
                                                            Date:  April 11, 1990
                                                            Page:  5
 1.1.4  Audit Equipment -
      The equipment needed for auditing is similar to the calibration equipment-
 however  the audit orifice transfer standard MUST be a different device from the
 one that is used for routine calibration and flow checks.

 !-2  Procurement Prerequisites— Laboratory Operations

 1.2.1   Filter Media -

      No commercially -available filter is  ideal  in all  respects.   The samolino
 Program should determine the relative importance of certa?n filter HZ   o
 criteria (e.g.,  physical and chemical  characteristics,  ease of handling cost)

       '
          lnlegriiy  " !5 ^9/m3  (a5s"min9  sampler's nominal 24-h air sample volume)
          measured as the concentration equivalent corresponding to the difference

          h±?6H thj initia  a"5 find1 W6ights of the fi"ter w^en weighed and
          handled under simulated sampling conditions (equilibration  initial

                                                              <™ -'s-Pler, re-
                     "/eSS th*n 0-005-mil1iequivalents/gram of filter as measured
                                   ' Re0ference 13 of the reference method (Subsec-
                                              '                                 and
                 Iilters may not be suitable for use with all  samplers
1-2.2  Filter Protection -
                                                            '


-------
                                                            Section  No.:  2.10.1
                                                            Date:  April  11,  1990
                                                            Page:  6


comparable size  (large enough to allow easy  removal of the  filtert_yet small enough
to prevent excess movement within the petri  dish) and have  a tight-fitting  lid to
prohibit damage  or  loss of particles during  transportation  to the analytical labor-
atory.  A label  can be affixed to the dish to allow proper  documentation when
sampling.  A sufficient number of petri dishes must be available to provide  (1)
protection for the  filter in transportation  to and from the monitoring 'location,
and  (2) storage  of  an exposed filter for subsequent gravimetric or  chemical
analysis.

1.2.3  Laboratory Equipment -                       "         •

     The analytical balance must be suitable for weighing the type  and size  of the
dichotomous filters used.  The range and sensitivity are dependent  upon routine
tare weights and expected loadings.  A minimum sensitivity  of ±1 /
-------
Section No.:  2.10.1
Date:  April 11, -1990
Page:  7







{/)
UJ
1-1
_J
0.

V)
1

2
K
o.
JJ3
UJ
U.
o
1-
UJ •
5

•- 1
e o
1 C C
i — •
i C
• J
u







fl •
._
E
i.
•>>
a
u
u






JC
u
jc !
u
1 u
c
u
u


1


41
c
1

._
3
o-
UJ







ti
• •— 1
i s
• L
• o
-
U M
• 3
e ?
a: <



c 4
•" *-
-o a
J-J 5
— * 'I
-tj .- ,r
3-0 —
o c
• X •
• a. o *>
J< II
4>' » O
• S U C
U 10 • .-
•'+• a: c -o
— u. ~ o
O LJ _jff ^*
S.S "^
VI T * *
• si
0 -0 J= •
c c -u —
f«A
— c •
5.?^ i
o S^ i
c -o • i
> 1-
• • "S S3
"•"of Z
I3 o 1
o . ° .f .
0 • • • «
a u a-o
iSS •!:
— • <- jc -o
Q-"o e o c
(O O t. 0 «

«
c
s °
I s s
•j = 8.
"• Oo
i- 'Jj 2 ^
• u • •- u
g- « L. — 0
0 — J3 Q. »)
_ Q- — a. •
^6 — 3 (.
~ • • «l O
"to • d j3
= J





L
~

J
•™>
a;

, .
• L.

•~ O Qi
•o *"* •
S-;"
— —
3 -O -O
O C —
e >
• a.—
c a-o
0 < C
• sTjc
U ID *>
? « '*
'o u •
as LD


t
• i.
jj O
ii A
rfxtwnts. 1
nd suitabli
'3
• • d>
L a c
• E-
*> • a
e -o 6
• CM
2 3 fl







L.
._
U.



L,
O L.
O
fl «•
• -D
™ J=
"g I
•
4» >
U O
• L.
••"I &
K •-


jj
•
C O
*^
— u
«- Q.
E *-
I «
*C fl
u ^
o c
«•• •
;&
= 1
< -o






O9
|
•o
o
•
u
'C
•
•o
>
e
o



a
c
c n

JC «
u u
e «
•*> -u
«— ?
^


fl
u
«
*>
a
3
0
a
ti
c
•
i





































J
"?
u
o

3
.^
•e



i
i.
a
»
—
—
x •
u -u
• X
<- e
3 -U
U
u c
• —
c -o
— 0
4> C
X •


t

v>
z
st ASTM or
tandards.
c •
• o
a—
• J3
•X f)
U 0
• 9
jr i.
it ^j

4> C
C 0
13
a. •
3 C

-------

-------
                                                           Section No.:  2.10.2
                                                           Date:  April 11, 1990
                                                           Page:  1
2.10.2  CALIBRATION PROCEDURES

    This subsection presents the folio-ing aspects of calibration procedures:
         Basic  calibration  procedures,  calculations,  and  rotameter "set nnint"
         adjustments  for the  dichotomous  sampler (Subsection  2 2)[

         Recommended  transfer standards and  calibration equipment  (Tables  2.1 and


         Sampler calibration  frequency requirements  (Subsection 2.3).


                                             requirements and calibration proce-
                 F1ow-Rate Measurement anrl
                                                     r consists of .... „_ UU01U

-------
Section No.:  2.10.2
Date:  April 11, 1990
Page:  2
bration of
t*nd*rd reference
M — •
ii k* e
*m



VI '1 •>
UJ |[ 1

tc < •
I-;
iy i

LL •)
•*• "1 *
M :
i! e u
i ii e •
5 !! .".a
•- •» e
o « o
s J 4) e
1 ,[ z S


S ii
£ * '

S •]
UJ 1 -
§ ii

5 i •
< ij *i

! 'l 5
§ t
s !!
i *
v) i!

a i!.


i •!
UJ 6 **
«c c
1
^ « "^
1* UI
i i
*~ 

;i « or
, i » •
3-?
• O k






•
k -O
• k
• ^
c c

k -*
1

•1 Protection Agency. *
•librating e Ltmintr Flow
• in.t «n N9S Ctlibnted
pereting Procedure*.
-M3, November 1801.
4* kf CIO £>
if O -^ k 1
e •• uj • a.
O Ik T3 O
k • ^ C (O

> k *> a.
C 3 *> V> >-

v> o e uj M
3 a. ui -i ui






or

II
^.^
u.
u


ex


••*

1 -0
^B k •
e • e 3 e
k • • Q. k
e ut 3 f e 3
** u. *> • ja *> e
ZJ S" .^ 2 -S
». o— o • • e k

>*k o e c e —
•ecu* • «> e e
— e c e E — a







e
1
k
£
0 f
e •
1 | S3
ui e c — •
U. JK • •- -0
-j •- a u. <





e

0)
•" C
ex _i




^^
*>
k C


'I —
• e
M

O
Uj *
u. •»•
— *
P-:
• ex
k •->
— 1-1 e
•c o
•s '"s
o S * **

3 S O f*
•• k • F*
•i a o. i
< « V) V

^<-i
^ e
• o *» <
D — C 0_
C7 •** • IU



3

cr

ii
^^
u.



i


*•"*

.'-
i S o •
15 fe e .2
^ *> k *9
3 J3 C
•n o — o
e — «
^ o "3 *>
i * *
ellt
e aa "





k
§

1.
f
1 • k
Ske
c e *»
k *l O.
a « — •
U, -e — -o
a i- u. <







o> e
7 'I
e -~.




a
o



•
Z^n^
t.
*^ •

2 Z
3
3 "
^ < n
k «-T
» 1^ C
< ..£ _
k — ' 3 f^
o o u *»
*• *J!S
•s • .*»
o 9 • M
"O •' O Gl
C -i -C 3
o *» o r«-
f lit
* k e f*
• a a i


£• J.-8
» c
• o *> <
s — e a.
cr «> e ul



3
or*

H

,•*
JJ;
«_«
„


B, J

0


-
f
3 C
° -1
5 S3
0
.lit.
4* — e •
•o > J ?
•; £ S3
0 S.
V) C





J
1
k
J3
9*^
f 5 « k
5 • k e
t 9 • *>
k a** a.
3*0—0
O H- tf) U. <







Ok C
7 ^
c* J





•


k
-»•». n
«J M
a—*
6 k
*« X •
k *»
3 "D f
0 ^ C
O
Q Handbook for Air Pollu-

•
k
3
m

<

*»

•
cr
i
4


M
JO
»

M

•^
»•"-•
•
Q_ ».


J^


e



o
c
k
c
X
o
o
s




1






u
*>.
k
o







1
O4










0
u


"k
o
i«
-c*
MI e
. *£
— *>
i **•
»o
•1 « •»
jj K
• •> a

res
ISR
3 a i
• V) V
• -^
j^|

0 *> <
•~ jb &
•*» • UJ
y
f







1
II


4

rf








' C
«

^




9
k
1
k
<
*•
<•> k
15 ik
If i S
k O ••* O.
• c — •
X • — -D
H- 3 U. <







e
'^



















i
t



i
i

'
t
'.
i
l

i




: j
. 0
1 *.

i k



i :

i X
i t
! *>
1
i ••
• .*»
 C

k
1-
•










































C
o
5
c
o
u
k
•w
c
•

a.
UI

o

•5
u
S
k
•
*j »
k
.^ '
c
• *

•

£
0
U

£
J3
i
o
u
Ik
a

41
C
O



k
e
o
k
"•
u

a
•«
c


•
k

^;
c

. k


JC
•*>

5
3

C

*
.k
a
*
JC

S
k


k
>
X

•
k
o

a
c
*c
£"~






0
Tl ^> C
** *» 0
«, • —
.** K - "j
3
" c ^ J
»- a.
' c
o

^» •
or k k
J> O
H — *J
— U
• • «
or w •*.
u
*
•
•
k
•
o
<*»
«
•
II
•
a

rofiwtrl
J3


e
k
3
*


£
X

£

k
3
U
U
•

O

•
&
u
k

J
o
•
•
*l

*>
m

k
e
e
•

2
Q
k
3
k
!
O
C

k
3
•

!
e*.
O
•

J3
Ok
•
u




2
K
g
k
•

-o
U
Ls
«
J .
**
i
c
o*
*>
o
«
«
k

!
i
«


c
o

«>
3
O
ID
J

C
.c
*>

•w
c

o
e
a
V
u
«
O CX

a
k
.
3
|
c
J!

o .
j=


c
s
0 *

o
Z!
"S
IJ
ll-
k C
0 •
•
UJ •
U» k
-j a

9 k
X
** *9
C
o
K

.- *.
a o
k •
o a
c
c *

— o
o
a v
•o o

9 •»
X
•









.


2



k
j
c
o


II

3
C
"i

^
•
s

s
O
II
^
k

C


I
*>
a
c •
k
3
*
2
X
3
U
U



o


M
&
u

k
j

**
^
o

x
u
^
t
o.
o
v»


-------
Section No.:  2.10.2
Date:  April II,  1990
Page:  3

1


1

i
!

j

}

1
I
U)
.IL '
UJ
tn

o
u

i
2
3 ' '


Hi
a >
2 ""'
lO . i
rANDAROS AND A!
1/1

'o
UJ
2

S
o
s.

ui
<
*~ •:







E
£

«»
a
o






— •o
, "• •
x u « . .
O • L • •*
! c e -o • oo
• y O°C *~ 3
S i * .3 J "S u
i c o •
• i e • in o .0
~. -«5ii
• •"• s .s = » s
C T5 . *> »> • e 2
0 U 0 • — *
— • b t. «> 4* .
•> -o ; tt. .0 • •  w oo a
• i. ' i u<-0§
w e i I o ~ L i
*• c •»• uj • a.
• 0 U. T> 0
c L m _i e «
e — » — •
i. ; > L -ja.
*» i e r *i (/> i-
i tu *Q e a;
!«-H«d
• L — u. 3
1 3 a. ui _i u

!
i

c u *»
0 . . m
— A i or
* e •
e o n
•a 'Z , ^
•"• • . u,
— 3 W
• 0- i ••-•
u • .! ^

• ; *s

i 1 *O
i • '•• i 1 .
e *" c Z ^ S 5 ?
' "a i. * c *

* • 3 '^ 0 • _
, in v> a
,
• • ^
' x
: u ft e
•» s -o
fc
i. 3
»> ° • •*
1 ^ I W
.. >• e •
3 , ** J "•
°* • U
u f e • u •
o c e L
1 E — - *> •
1 >. — • O.Z
' UI « u C • —
( U, .c ^ — -0 .-
1




c
or 'i
• ' ^
• 0
O C CM

^ L *•

' L C
s? . si
•: . ~c *
••
t- « UJ —
U. 0
2~ 3 i
— f . Z! j/>
v <•(/>?
^5 — c e 5.
S .2 2 5 S .2 .2 8
CT *> • uj or«>«i«

T,
•» •
•S or


« ^
u.
u. ^


*"• •
3 f |
> ^—

i • •
• 3 0 • -C
— . •" = t» o
• o • - 3 c
•tf a o «> "Z 2 • 'i
j> 3 i ? ** § **
— "S°^g f^'S
* ix~ — s^i--
j^j §^^I-§ sue<»

a - e
• _o u s> o
or 3 < « ^
3
n


^,
• y

C II
° 1 "°
o u> ^
I i=
"Si
fc c ^J
^/ M

i i
I :J

*» S "o
1 IJ
§0 •
.- *>
s 1 1
.ill
8 -° J .-

^* t 1 L •— "

M i
' ^ *-^ 1
7—^ . fc
a. H-

O a 1
^L U
« i.:" ' i
^ 3 C
C • o *>
* 1 ? 5
> 1- X "S •
0 • — • c

t 1 • : s
• • ' i = -? •
i : s: 5
ii S 5 ^ S J5
- . !!• • «• :r--
i 1 ili: ^ Is I
i. i.illl 1 il,i
• • — : : — « ISN"
o ^ o — — u y ' ! *" ' ^ « • ™

0 3o3o».'— ^ .|• a
ij *. *• "S • • 3
•". ' ^ • iT S e~ '•
• .• jij >!?^?
i a ii '
= « 1s • E a "
• g 6 -c • y o
s ~ • • - a 5 .5

• • ^' '1 S 1 " »r'5S
• •13 S^« i'^° * 5*L*
C'§ I •» * • 'if. = ' i .5 J •;
ot-a «uia.u. !»3 7 -»o*3
. ; • • .. *> y
jiS< E ?oaS


-5 .5 :' I o ' -S 3 i i «
•1 E i:c*' ,.st?»



~ 7 '25 - — . • "s^«?
- *> j i i J 3' 3 § »°K
• • "5 u 3 a • ^ .-
'• 5 "e Vs.* ^ u u a *"'

i. ! • * _ o e c •
-3 '•'• ^ ? - f -?» -5
J c| i:-i-;«»">-'ia««
•^ 3a ' ' • c o ii ~5c"°-a


-------
Section No.:
Date:  April
Page:  4
 2.10.2
11,  1990
























^M
35
QJ

UJ
cc
3
UJ
ce

z
2
[j
ce
co •

^
0
t
0.

UJ
CN
UJ
CD
<^
t—

i




















1 49
?!

3 ^
cr o
• c
k
9
«•- k
•~ •>







•
0
*>
^
t.
.O
C • 1 •—
0 49

3 •
u 6



J5
•*
E-



0
u
c
.3
o.
u
u











c
T,i
C k
• 3
X •
u 0
i
3 «•-
' cr o
0
k TJ
U. 0
jc
41
I
1
i



— " -w
£
u
cc
It

_3 K
• V
k +1


O 41
m"t A
S. k

TJ *
• t-
u

TJ 3
C k
l-l 4>

^
TJ k • •
k • 4> 49

t"> *"
• k
49 • TJ O
II —
e • •». •
k o jc •
• U T> 49
• U -C 41 k
ecu*
• • u k
k c k — o
49 • 0 T> —
4» — c *•
J*~7~c
3 k « •
k • 41 E TJ
0 49 «— o
U «• TJ k 0
• 3 fc jc
5 . * x **
.- 49 0 i- £
K o. e »i o
4i u •
• 0 XJC K
k k C 4» ta
^3 H **4
— C • +1
— O T> k
• 0. C 0 k
«J 3 • I O

* I
( ' 1 . , .




*> 1
c
Q.
•—
3
UJ




k
1
•
VJ


1
k
JO

^_
•
o •

k >
• k
C 3
u

Q. e
0 0
5 3
;5
k k
.Sik
• — o
k • —
u «-

0 • —
— 3 •
•fc O 3

TJ > U
41 k
• 0. II
u
— E C •
TJ 0 0 X
C k — CM


1


k U
X • *
1 k 49
*> -iT-S
• k —
4» TJ •
• c
• *»-.
~ * •
49 >, d
• k •
ill
4> ^i
Ck4>
•^ • 0
9 C U
• "• —
t. a a
C k
0 «
a— k
3*0
JC k •
o • *>
-C C L.

k
o

n
c
i

*> TJ
• k
k •
k c
o •
••• 4)
u. «


k

^
g^
4)
k

k
o


p
3
T?








•
JC


M
c
1

e
m



,

1

e
o
x

c
3
o
k
c

0
II
^
JC •
u •
k X
3 •
a-o
49 X
• k
JC >
u o
• u
U k






k
1

4>
^
0
c
o


• 0 •
U 49 U
• c
— k «l
CL O 49
• 4> a. •
k • «
u u
k — O
o TJ e
c
49 — C II

?C •"
-0 .- 43 .S
•< 4» • —








•
JC

CM
c
"i

CM



c
o
c
Jf
"•- JC
O 4>
i!
IX
I-
4» TJ
e
TJ •
k
TJ Q.

4> U'
II •
k
c
J= O
4> 0.
— 3 •
X —
• 0 •
k « >
• k k

o u c
(^ • *^






1

1
TJ
o <•
H 1.
a «


•o
U TJ
-4?
a •
e k
k J3

• ^
o •
c u
'H 0
«• ^
10
JB k
o
'• *»»
> TJ
* c
X •


» 0
™^L ^* oj
i^r
t* CM

III »l

X II
4> II
... ...a
> x aua
— o E o
... k S C
IH 3 <-l N
e u

v> < • •


IS
e
a • a
• c
« k
j?.T5
U JC O
k U U
S • •
O.JC • •
U 49 4>
ii c <3 a
o — —
a— • •
c 4> k k
j: k TJ k
13 J3 k •
"l^i^
1 O • •-

(> U • 0-
t«- 0
4* ^0 9
0 > •
u *— — o
— t_ •». c
k • •
49 a. o k
!!»"-
~ e • •
> — • E
a k k k
k 3 j: o
O TJ 49 C

0
u
c
a

ii

3 • •
•-, U
TJ O C
< 4> •

a
II C
09 TJ
e •

TJ k

• k


ll
• k
U JC
— u •
TJ XK
C • CO
M a »i




c
f I
3 1
J3 CO
X4>
TJ
49 C
*- a
o —
• u
a •
• e k
TJ C
• o

«• =
-C k
* i —
• 0 •
k k >
• JC k
cx u •
E X4>
one
u a —


x

.^
TJ
'I
JC

• O

4> U
• c
K —


























•

U
c
~


^_
0
u
"x
—
c

JC

o
.g
k
• c
k •
TJ •
k
O —
u •
^ ^
0
• 49
41 C
^ S
o •
«•- n
• u
3 0
0 -C
3 U
C.
C CO
O —
U •
k

OS
•C Jl
— S
k *^

9 C
C •
— 0
41 k
Ok 0

• -O

-------
                                                            Section No.:  2.10.2
                                                            Date:  April  11,  1990
                                                            Page:  5
      As indicated above, the true or actual flow through the sampler inlet must be
 known and controlled to ensure that only those particles nominally less than 10 w
 are being collected.  A common source of error in a PM10 monitoring program is
 confusion between various air measurement units.  Although the sampler must operate
 in terms of actual flow rate (Qa) ,  flow rates must be corrected to EPA standard
 h^hn10nS^°o ^mP?rature and Pressure (Qstd) before data can be submitted.  Thus
 both Qa and Qstd flow rates are used for PM10 measurements.  Before calibration
 procedures are initiated, the operating agency personnel should- review the
 following flow-rate designations:

      Qa:  Actual volumetric air flow rates that are measured and expressed at
      existing conditions of temperature and pressure are denoted by Qa
      (Qactual).   Typical units  are  L/min and m3/min.  Inlet design flow rates
      are always  given in actual  volumetric flow units.

      Qstd:   Air  flow rates  that  have been  corrected to  EPA standard conditions
      of  temperature and  pressure (25'C  or  298 K,  and 760 'mm Hg  or 101  kPa)  are
      denoted by  Qstd (Qstandard).   Typical  units  are std.  L/min  and std
      iii-Vim n.   Standard volume flow  rates are often  used  by engineers and
      scientists  and  are  equivalent  to mass  flow units.   Standard  volumes
      (derived- from standard  volume  flow rates)  are  also  required  to be  used  in
      the calculation of  mass  concentration  (^g/std.  m3j  in  reportina PM10
      measurements.                                                   ,

      These Qa  and  Qstd measurement  units must not be confused or  interchanged   The
^/ate,units  can  be Converted  as  follows,  provided the  existing  temperature  and
periodT  arerknown?me  "^    * 3Verage  temPerat^e  and Pressure o?er a'sam^ing
                            Qstd = Qa(Pa7Pstd)(Tstd/Ta)                      (Eq. jj

                            Qstd = Qa(Pav/Pstd) (Tstd/Tav)                  (Eq. la)

                  . •   •        Qa = Qstd(Ta/Pa)(Pstd/Tstd)               -     (Eq. 2)

where:

     Qstd = standard volume flow rate, std. m3/min
       Qa = actual volume flow rate,  actual m3/min
       Pa = ambient barometric pressure,  mm Hg (or kPa)
     Pstd = EPA standard barometric pressure,  760 mm Hg (or 101 kPa)
     Tstd = EPA standard temperature,  298 K
       Ta = ambient temperature,  K  (K  = °C + 273)

     QTtd - average standard volume flow  rate  for the sample period, std.  m3/min

       Qa = average -actual  volume flow rate for the sample period  m3/min
      Pav = average ambient barometric pressure during the sample'period,  mm  Hg (or

      Tav = average  ambient temperature during  the sample  period,  K.

-------
                                                           Section No.:
                                                           Date:  April
                                                           Page:  6
 2.10.2
11,  1990
2.2  Sampler Calibration
     This subsection presents flow-rate calibration procedures for the most common,
commercially available dichotomous samplers.  Calibration procedures may have to be
adapted for other sampler models.

     The dichotomous sampler operates at a total actual flow rate of 16.7 L/min.
To ensure correct fractionation of particles at the inlet, this flow rate must be
maintained within *10% of 16.7 L/min.  The coarse flow rate is approximately 10% of
the total, or 1.67 actual L/min.  it must also be maintained to ensure correct
fractionation within the sampler's secondary separation system.

     Accurate calibration data for each dichotomous sampler are essential for the
following:

     1.   To determine sampler flow rate set points.

     2.   To establish sampler flow rate control limits.

     3..   To calculate sampler flow rate during routine QC field flow checks and QA
          performance audits.

     4. .  To calculate to.tal sample volume for the computation of PM10 mass
        "  concentrations.

Calibration of the sampler rotameters must be traceable to NIST standards.  A
primary standard is used to calibrate a transfer standard, which in turn is used to
calibrate the sampler rotameters.

     Several commercially available transfer standards can be used in calibrations.
Tables 2.1 and 2.2 list recommended standards, their applicable flow ranges, refer-
ences for transfer standard calibration procedures, and the equipment necessary to
perform sampler calibrations.  The following are essential considerations in choos-
ing a transfer standard for subsequent rotameter calibrations.

     1.   The transfer standard must be traceable to NIST through the calibration
          procedures referenced.

     2.   The transfer standard must be calibrated in the appropriate flow range.
          A minimum range of 12 to 19 L/min (total) and 1.4 to 1.9 L/min (coarse)
          is recommended.

          Note;  If the transfer standard has been calibrated in terms of EPA
          reference conditions, indicated flow rates 'for each rotameter setting
          must be corrected to actual flow rates (Qa) to determine the sampler's
          set point.

     As indicated in Tables 2.1 and 2.2, each transfer standard has a certain
value.  The operating agency should carefully choose the method that best utilizes
equipment on hand and minimizes difficulties in establishing traceabi1ity.

-------
                                                            Section  No.:   2.10.2
                                                            Date:  April  11,  1990
                                                            Page:  7


      Regardless of the transfer standard employed, a leak-tight adaptive  device
must  be used to connect the transfer standard to the sampler inlet".  Figure  2.1
illustrates such an adapter.  These may be purchased comrnercially or fabricated
in-house.  Obviously, the corresponding outlet on the transfer standard will
determine whether a pipe thread or tube fitting will be attached.

      Tables 2.1 and 2.2 present only the basic apparatus necessary to perform
calibrations.  In addition to those listed, the operator will need a few miscellan-
eous  supplies. _These include a 9.53-mm (3/8-in.) Swagelok cap, 6.35-mm (1/4-in )
Swagelok cap, and hand tools.

      A station log book or calibration data sheet must be used to document
calibration information.   This information includes,  but is not limited to
instrument and transfer standard model  and serial numbers,  transfer standard
traceability and calibration information,  ambient temperature and pressure
conditions,  and the collected calibration  data (rotameter units versus  indicated
flow  rate).

2.2.1  Precalibration System Check -

     Procedures for the precalibration  system check  are  as  follows:

     1.    Place a  pair of  filters  into  the  dichotomous sampler  filter holders
          Filters  used for flow  rate  calibrations should not be used for  subsequent
          sampling.               .                                      .       ^

     2.    Remove the  sampler's inlet.   Turn on the sampler  and  allow it to warm  up
          to  full  operating  temperature (at least  5 min).

     3.    While the sampler  is energized, slowly  close off  the  inlet tube  with a
          rubber stopper or  duct tape and observe the total  vacuum gauge.   If the
          sampler  is equipped with an overload feature,  it  should shut  down  the
          system when  approximately 15  in. of vacuum is  reached.

     4.    If the sampler is  equipped with the overload feature, disconnect    Next
          perform a system leak check by opening both rotameters completely  and ' '
          sealing the  inlet  tube with a rubber stopper or duct  tape.  When a  maxi-
         mum indication on.the total vacuum gauge is reached,  shut off power to
         tne unit  record the maximum  reading on a data sheet, and observe the
         rate of decline in the readings of the vacuum gauges.

         Note:  Leak-free systems should indicate a vacuum of  10 to 15 in  or
         more, and the rate of decline to 0 in.  indication should require 60 s or
         more.  If these conditions are not met  and the control module was
        •successfully leak-tested previously, a  leak exists either in the inter-
         connecting tubing or in the sample module.

    5.   If applicable, reconnect the overload feature.   Conduct a pump perform-
         ance check.   Open the inlet  tube  and apply  power to the unit  P When
         stable flow is achieved,  adjust, both rotameter  control valve  to 90% of
         the rotameter scale.  Observe  the total  vacuum  gauge indication.

-------
                                                     Section  No.:   2.10.2
                                                     Date:   April  11, 1990
                                                     Page:   8
    O-RIng
                                 Transfer Standard


                                       Noncrimp Tubing
                                       9.53 mm (3/8 in) NPT to
                                       Barb Connector
                                          •Place Teflon tape here
                       32mm —>j
                       38 mm 	•
                          To Inlet Tube
                                          .Stainless Steel
                                          or Aluminum
Figure 2.1.  Inlet adapter that may be used to connect the
      transfer standard to the sampler's inlet tut*.

-------
                                                           Section No. :  2.10.2
                                                           Date:  April 11, 1990
                                                           Page:  9


          Note:  Consult manufacturer's instruction manual for minimum vacuum
          indication.  Readings lower than specified vacuum readings indicate
          possible pump diaphragm or reedvalve problems, which should be
          investigated and corrected before continuing with the calibration.

2.2.2  Total Rotameter Calibration -

     Procedures for calibrating the total  rotameter are as follows:

     1.    Set up calibration system as illustrated in Figure  2.2.   The inlet of the
          transfer standard is open to the ambient air; the outlet  of the  transfer
          standard is connected to the inlet  tube of the dichotomous  sampler.

     2.    Turn on the sampler and  allow it to warm up to normal  operating  tempera-
          ture (at least  5 min).   If an electronic transfer standard  is  used  it
          must also equilibrate before proceeding with  the calibration.

     3.    Adjust  the total  flow control  valve to  approximately 90% of the  rotameter

          of 167 L/min       °°arSe f1°W C0ntro1  valve  to  indi«te a  nominal flow


     4.    Read  the  following  parameters  and record  them on  a data form (Fiaure  2 3)
          or in a log  book:                                               3        '

          •    Ambient temperature  (Ta) , K               •             .

              Barometric pressure  (Pa), mm Hg or kPa

              Transfer standard readings  (TS) , volts, AH20, timings,  etc.

              Sampler total  rotameter indication  (I), arbitrary units.

                 pr°5e5n.re !°r rotameter settings representing flow rates of 75
                         °f     establd operating range (12 to 19 L/min)   For

                                                          '
          each
                                                          '            "—ponding
2-2.3  Coarse Rotameter Calibration -


     1'    I?/S ?fVhe^Sa?p.1er-  disconnect the fine flow vacuum line [9.53-mm
           3/Sin'   ?:!L i!ej;  3?d Ca?-lhC fyVlow outlet-port with a 9.53-mm
           in?"™™ tn th      P ^^ Fl9Ure 2'4)'   Thl'S St6P ke6PS  the fine ^°«
          f if0^Ph0  ?+  ! vacuum pump.   It is recommended that a particle-free
          filter be attached to  the detached fine flow line to prevent particles

          sCnd^d     9 ^ SyStem'  In5td11 the C°arse f1ow ra'e transfer


     2'    h^h9?6  the samP1er and  the transfer standard (if electronic)    Allow
          both to warm up again  to  full  operating temperature.

-------
                                                Section  No.:
                                                Date:  April
                                                Page:  10
             2.10.2
            11,  1990
     Calibration
      Adaptive •
       Device
 Chosen
 Transfer
Standard
                              Coarse Flow
                                        0   0
                                Coarse
                                 Flow
                               Rotameter
  Total
  Row
Rotameter
Figure 2.2. Calibration assembly and dlchotomous sampler
           with transfer standard connected.

-------
                      Dlchotomous Sampler Calibration Data Sheet
                        Section No.:  2.10.2
                        Date:   April  11,  1990
                        Page:   11
Station Location Kalet'^k, , NC n^te *?[
Sampler Model 'Z'f'f £" -<$/N £,(Cj
Pa ^4? mmHq Ps ^l mm Hg
Leak check, maximum vacuum (^ in.
Total Trans. Std. Model M.FM S/N 2.113
Qa Cal. relationship m. 6-142 b. 0-oo
Coarse Trans. Std. Model SFFf^\ s/N V/<1
Qa Cal. relationship m . '• ° b - O

Cal. Indication Flow Rate
Point (TS) Cv
^ V* 0:1*1-3 \.bO
/3 3 - 
2-0 '(. O.3-I2. /.*fo <2 £- 6" Z^?'

                                                >^^^^^^^«^^i»^^Bi*H^HBBl^HMi^HiHi^ll^

Sampler Cal. Relationship (Qa, x-axis; corrected recorder response, AC, y-axis)
Total:
TFR
                                     CFR
Coarse: m « I

 X-^_ CSP
                             ' "   ^^^™» ^^ III  ^^mHB^Bo.
TFR = 16.7 (Ps/Pa)(TayTs)  CFR = 1.67 (Ps/Pa)(TaATs)
TSP, CSP - {[m (TFR, CFR) + b] [(Pa/Ta)'*]}


Operator _	


            Figure 2.3. Example dlchotomous sampler calibration data sheet.

-------
                                                             Section No.:
                                                             Date:   April
                                                             Page:   12
                                2.10.2
                               11,  1990
                   Calibration
                   Adaptive •
                    Device
               Chosen
               Transfer
               Standard
                        Fine Flow
Coarse Flow
Disconnect 9.53 mm (3/8 In)
O.D. tubing and install
Swageiok cap
    Paniculate filter
    Installed on
    9.53 mm (3/8 In)
    O.D. tubing
           0   0
                                                Coarse
                                                 Flow
                                              Rotameter
                       Total
                       Flow
                     Rotameter
              Figure 2.4.  Calibration assembly and dichotomous sampler
                     set up to calibrate the coarse flow rotameter.

-------
                                                            Section No.:  2.10.2
                                                            Date:  April 11, 1990
                                                            Page:  13


      3.   Adjust the coarse rotameter flow-control valve to an approximate value of
           90% of the rotameter scale.  Adjust the total flow control valve to
           indicate a nominal flow of 16.7 L/min.

      4.   Read the following parameters and record them on a data form (Figure 2.3)
           or in a log book:

           •    .Ambient temperature (Ta) if variation has occurred,  K

         .  •     Barometric pressure (Pa) if variation has occurred,  mm Hg  or kPa

                Transfer standard readings (TS),  volts,  AI^O,  timings,  etc'.

           •     Sampler coarse rotameter indication (I),  arbitrary units..

      5.   Repeat procedure for rotameter settings  representing  flow rates  of  75,
           60,  40,  and 20% of the established  operating  range  (1.4 to 1.9  L/min)!
           For  each calibration point,  record  the rotameter indication  and  corre-
           sponding transfer standard  output.

      6.    Turn  off sampler,  and  reconnect the fine flow line  and  the  sampler's
           inlet.                                                        v

 2-2.4 '"Calibration Calculations  -

      Gather  together  all  the  calibration  data, including the  transfer  standard
 calibration  information and  the  dichotomous sampler  calibration data sheet   The
 following  calibration calculation  procedures  are recommended.

      Note:   These  calculations should be  done  at the time  of  the  calibration
 rather  than  later.  This  approach  will  allow  additional calibration points to be
 taken if questions  arise  about the data that  have  already  been obtained.

     '1.    Verify that the transfer standard calibration equation  is current and
           traceable to an acceptable primary  standard.

   •   2.    Calculate Qa for each calibration point  as determined by the transfer
           standard calibration equation.

        •   Note:  It may be necessary to correct the indicated transfer standard
           flow rates from Qstd to Qa.   This can be accomplished by Equation 1.

                             Qa = Qstd(Ta/Pa)(Pstd/Tstd)          '          -(£q. 2)

where

             Qa = flow rate at actual  conditions   L/min
             Ta = ambient  temperature,  K (K = °C + 273)
             Pa = ambient  barometric  pressure,  mm Hg  or  kPa                •
     Pstd,  Tstd = standard barometric pressure  and temperature,  respectively.

-------
                                                            Section  No.:   2.10.2
                                                            Date:  April  11,  1990
                                                            Page:  14


      3.    Calculate  and  record  the  total  and  coarse  rotameter actual  corrections
           (AC)  for each  calibration point as:

                                  AC =  I(Ta/Pa)l/2                           (Eq.  3)

where            ,

      AC  s  actual correction
      I   s  rotameter  response, arbitrary units
      Ta  s  ambient temperature,  K
      Pa  s  ambient barometric pressure, mm Hg or  kPa.

      4.    On a  sheet of  graph paper,  plot the sampler corrected total  rotameter
           units  (y-axis) versus the corresponding calculated  transfer  standard
           total  flow rates (x-axis)  to obtain the dichotomous  sampler  total flow-
           rate  calibration relationship.

      5.    Repeat Step 4, plotting corrected coarse rotameter  units  vs. the corre-
           sponding calculated coarse  flow rates.

      Because the determination of the sampler's  average operational  flow rate (Qa)
during a sample period depends on the ambient average temperature and  pressure,   use
of a  graphic plot of the calibration  relationship is not recommended for subsequent      	
data  reduction.  This plot is used  only to visually assess the calibration points       |Bk
to see if  any should be  rerun.  •                                                        ^P'

      Plot  the regression line on the same graph paper as the calibration data.   For
the regression model  y = mx + b, let y =  AC = I(Ta/Pa)l/2 and x = Qa so that the
model is given by:

                          AC = m[Qa(transfer standard)]  + b                 (Eq. 4)

      Using a programmable calculator or a calculation data form,  determine the
linear regression slope  (m),  intercept (b), and correlation coefficient (r) and
record them on the  data sheet.   A five-point calibration should yield a regression
equation with a correlation coefficient of r > 0.990, with no point  deviating more
than 0.5 L/min for  total  or 0.05 L/min for coarse rotameter calibrations from the
value predicted by  the regression equation.  Plot the regression  line on the same
graph paper that has  the individual  calibration points.

  •   6.    For subsequent  sample  periods,  the sampler's average actual  operational  •
          flow rate  TQa or CQa  is calculated from the calibration  slope and
          intercept  using Equation 5:                                  •         .


                        TQa or CQa = l/m[T(Tav/Pav)1/2 . 5]                  (Eq. 5)

-------
                                                            Section No.:  2.10.2
                                                            Date:   April 11,  1990
                                                            Page:-  15
 where
      TQa,  CQa = sampler total  or coarse average flow rate,  actual  L/min
             I = average total  or coarse rotameter response,  arbitrary units
           Tav - average ambient  temperature for the  .run  day,  K
           Pav = average ambient  barometric  pressure  for  the run day,  mm Hg or kPa
             m = slope  of the total  or coarse flow-rate calibration relationship
             b = intercept of the total  or coarse flow-rate  calibration relation-
                 ship.

           Note;   The expression  [T(Tav/Pav).l/2]  is the "y"  term of linear regres-
           sion  equation:   y  =  mx +  b,  or x  = (y-b)/m.

           Note:   Tav and  Pav readings  may be recorded  on-site or from a nearby U.S.
           National Weather Service  station  or airport  weather station.   Barometric
           pressure readings  obtained  from remote sources must be at station  pres-
           sure  (not corrected  to sea  level),  and they  may have to  be  corrected for
           differences  between  the elevation  of the monitoring site and  that  of the
           airport.  If ambient temperature  and pressure  readings are  not  available,
           seasonal average temperature  (Ts)  and  barometric  pressure (Ps)  can  also
           be  used.  Care  must  be taken,  however,  that  the actual conditions  at the
           site.can be  reasonably represented  by  such averages.   It is  therefore
           recommended  that seasonal values  represent actual values within  20  °C and
           40  mm  Hg.

2.2.5  Rotameter  Set Point Adjustment  Procedure  -

      1.    Calculate and record on the calibration data sheet  the total  and coarse
           seasonal flow rates.   These values  will be used to  determine  the seasonal
           set points for  both  rotameters.

                              TFR = 16.7(Ps/Pa)(Ta/Ts)                       (Eq. 6)

where

        TFR = total flow  rate for adjustment  of  the sampler total  rotameter
       16.7 = design flow  rate as specified by the manufacturer, L/min
     Ps, Pa = seasonal  average and ambient barometric pressure,  respectively   mm Ha
            .  or kPa  .
     Ts, Ta = seasonal  average and ambient temperature, respectively,  K.

                              CFR = 1.67(Ps/Pa)(Ta/Ts)                       (Eq.  7)
where
        CFR = coarse flow rate for adjustment of the sampler coarse rotameter
       1.67 = design flow rate as specified by the manufacturer, L/min
     Ps,  Pa = seasonal average or ambient barometric pressure, respectively, mm Hg
              Or Kr3
     Ts,  Ta = seasonal average or ambient temperature,  respectively, K.

-------
                                                            Section  No.:   2.10.2
                                                            Date:  April  11,  1990
                                                           •Page:  16


      2.    Calculate  and  record  on  the sampler's  calibration data  sheet  the  set
           point  rotameter responses  that  correspond  to  TFR  and  CFR"  calculated in
           Step  1.

                            TSP  = {[m(TFR)  +  b](Pa/Ta)1/2}                    (Eq. 8)

where

      TSP  s total rotameten set  point,  arbitrary  units
      TFR  s total flow  rate,  L/min
      Pa  s ambient barometric pressure, mm Hg or kPa
      Ta  s ambient temperature, K
       m  - slope of  the  total flow-rate calibration  relationship
       b  s intercept of  the total  flow-rate  calibration relationship.

                           "CSP  = {[m(CFR)  +  b](Pa/Ta)1/2}                    (Eq. 9)

where

      CSP  = coarse rotameter set point, arbitrary units .
      CFR  s coarse flow rate, L/min
      Pa  - ambient barometric pressure, mm Hg or kPa
      Ta  = ambient temperature, K
       nr-s slope of  the  coarse  flow-rate calibration relationship .
       b  * intercept of  the'coarse flow-rate calibration relationship.

      Adjusting the sampler  rotameter  to seasonal  average conditions will  help
minimize  data loss caused by exceeding the manufacturer's design condition
specifications.

      1.    Energize the sampler  and allow it to warm up to operating temperature (3
           to 5 min).

      2.    Following the  manufacturer's instructions, adjust the total rotameter
           until the sampler response  indicates the total flow-rate set point (TSR)
           as calculated  in Step 2 above.

      3.    Following the  manufacturer's instructions, adjust the coarse rotameter
           until the sampler response  indicates the coarse flow-rate set point (CSP)
           as calculated  in Step 2 above.

      4.    Verify'that the sampler will maintain  these flow  rates for at least 10
          min.  Turn off the sampler.

      5.   The sampler can now be prepared  for the next sample run day.

2.3  Sampler Calibration Frequency

     To ensure accurate measurement of the PM10 concentrations,  calibrate the samp-
ler upon  installation and then recalibrate it as follows:

-------
                                                      Section No.:  2.10.2
                                                      Date:  April  11,  1990
                                                      Page:  17


1.   At least annually.

2.   After any repairs that might affect sampler calibration.

3.   If the field calibration flow check results exceed QC limits (+10% from
     the sampler's required design condition flow rate or ±7% from the
     sampler's indicated flow rate).

4.   Whenever an audit indicates that the sampler is out of calibration (±105
     from the sampler's required design condition flow rate or ±7% from the
     sampler's indicated flow rate).

-------

-------
                                                            Section No.:  2.10.3
                                                            Date:  April 11, 1990
                                                            Page:  1
 2.10.3  FIELD OPERATIONS

 3.1  Siting Requirements
      As with any type of air monitoring study in which sample data are used to draw
 conclusions about a general population, the validity of the conclusions depends on
 the representativeness of the sample data.  Therefore, the primary goal of a PM10
 monitoring project is to select a site where the collected sample mass is repre-
 sentative of the monitoring area.                      .

     .Spatial' and temporal scale considerations are important in dichotomous sampler
 siting. .Spatial scales may range from a small (0.1- to 0.5-square kilometer)  area
 to large regional areas exceeding tens of hundreds of square kilometers.   Whether
 the potential  impact of particulate pollution is generated by a local  or  qeneral
 source category will  affect the decision on the size of the spatial  monitoring
 ™?i   *'  ! ?S    !?"'    e S1'i1f)9 Of the samPlers W1'thin a  monitoring  .network should
 reflect whether the expected impact will be limited to a  small  area  (a few city
 blocks) or extend to larger areas (metropolitan or rural).

    .With  regard to the temporal  scale,  interest focuses  on either an  annual geo-
 metric mean concentration or a  24-h average concentration.   Because  siting of  a
 dichotomous sampler requires that consideration  be given  to prevailing wind direc-
 tion,  a sampler sited  for monitoring trends in air quality  over a  period  of a  vear
 of l^r^r^n6 JdeaT *>r ^asuring 24-h concentrations.   Thus, the choice
 of temporal  scale will  also affect the.sampler location.

     Although spatial  and temporal  scales  must be considered  in  site selection  the
 following  guidelines should be  observed  regardless  of  the scale:       cie"lon, me

     1.    The dichotomous sampler must  have unobstructed  air  flow  for  a minimum of
           2 m m  all directions.

     2.    The sampler  inlet  should  be .placed at  a height of 2 to 15 m  above ground
     3.    If a dichotomous sampler is collocated with any other particulate sam-
           pler, the minimum spacing between sampler inlets must be 2 m and the
           maximum spacing must be 4m.  All inlet heights should be within 1
           vertical meter of one another.

     Complete siting requirements are outlined in 40 CFR 58, Appendix E.

win £IdHt1?nayaCJ!!rS mustnbe considered in determining where the actual sampler
ab  ifv ofPl2n ;t  h?SV"0 Ude acceS51'bili*y under all weather conditions  aJa^
ability of adequate electricity,  and security of the monitoring equipment.

     A dichotomous sampler used for routine sampling must be situated where thP
operator can reach it safely regardless  of weather Editions!  if the sampler is
located on a rooftop,  care should be taken that the operator'^ personal  safety is

-------
                                                           Section No.:  2.10.3
                                                           Date:  April  11, 1990
                                                           Page:  2


uwb jcupuiuitcu uj u ji.p^ij i «w i Ju, ,uv.v. uw. ,..y  . ..w . w...w.... ,.^-w..w. .  ».„...,.«_.-
tions also should be given to the fact that routine operation  (i.e~., calibrations,
sample installation and recovery, flow checks, and audits) involves transporting
supplies and equipment to and from the monitoring site.

     A dichotomous sampler will require a minimum continuous operating current of 3
to 5 A (120 V a.c., 60 Hz) and may require a higher startup current, which necessi-
tates a slow-blow fuse.  Although most dichotomous samplers are equipped with
timers, there is often no recording device provided to indicate short-term power
interruptions.  This lack necessitates a stable power source for the monitoring
site.

     The security of the sampler itself depends largely on its location.  Rooftop
sites with locked access and ground-level sites with fences are common.  In all
cases, the security of the operating personnel as well as the sampler should be
considered.

3.2  Sampler Installation Procedures

     1.   On receipt of a dichotomous sampler from the manufacturer,  visually
          inspect the sampler to ensure that all components are accounted for.
          Compare equipment delivered with the enclosed packing slip.  Notify the
          manufacturer immediately of any missing or damaged 'equipment.

     2.   Before transporting the sampler to the field site,  perform a quick labor-
          atory check'to determine if the sampler i.s operational. • Energize the
          sampler and observe rotameter responses, vacuum gauges, and pump perform-
          ance.

     3.   Carefully transport the sampler to the monitoring site.

     4.   Bolt down the sampling module to a secure mounting surface.

     5.   Install the control module.  This module can be bolted down adjacent to
          the sampling module (no closer than 2m), or it can be located remotely
          (e.g., inside a monitoring station).  It is recommended that the control
          module be no more than 10 to 15 m away from the sampling module to avoid
          a pressure drop along  the flow lines.

     6.   Connect the vacuum lines between the sampling module and the control
          module.  First,  hand-tighten the nuts on the tube connectors as much as
          possible,  and then wrench-tighten them 1-1/4 revolutions.  Be careful not
          to cross-thread the fittings.

     7.   Check all  tubing for crimps, cracks, or breaks,

     8.   Plug the power cord into a line voltage outlet.  The use of waterproof
          interlocking electrical  connectors is recommended to ensure operator
          safety and to avoid shorts and/or power interruptions.   Do not allow any

-------
                                                           Section No.:  2.10.3
                                                           Date:  April 11, 1990
                                                           Page:  3
          electrical connections to be submerged during periods of inclement
          weather.

     9.   Perform a multipoint flow-rate calibration as described in Subsection 2.

3.3  Samp 1 i nq Operat i ons

     Sampling operations provided here are specific to one type of commercially
available dichotomous sampler.  Because operational procedures may vary among samp
ler models, the manufacturer's instrument manual should be consulted before the
sampler is put into operation.  Sampling procedure checks are summarized in Table
J » i •

3.3.1  Filter Installation


,rp -in^rr-136 ^kel! t0 enSUre *?at Clean f i^ers are not damaged  before they
are installed in a dichotomous sampler.   Filter cassettes should be  kept in their
protective petri  dishes (see Figure 4.1),  and any damaged filters  must  be
discarded.


     Thr Pr°5?dhr? US6d t0 Tta11  fi1ters  in a dichotomous  sampler  is  presented
     ch  ?5  dichotomous sampler is  equipped with two filter  holders,  and the  petri
      nn   h   Hmari?dKt0 ind,i"te which  filter will  be  used  for  coarse  particle
        andh;ne  sam^nYrun tll^*  '""«<*"»"<*•  the filter ^sampler ID
     1.    Switch  mechanical  or digital  timer  to  "OFF."

     2.    The  coarse-particle  filter  holder is the one with the 5,35-mm. (1/4-in.)
          o.d.  tubing   and the Tine-particle  filter holder is the one with the
          erfrZ iun~in'H-°;      S%\ As*hown in Fl"9ure 3.1, the filter hold-
          ers  can also  be distinguished by the fact that the coarse-particle filter
          holder  is on  the center  line of the virtual impactor head and aerosol
          l*n%'£ ^r635/^ fine-particle filter holder is offset.  Unscrew (by
          a^eLlJ6 k"urje
-------
Section No.:
Date:  April
Page:  4
 2.10.3
11, 1990




















10
UJ
X

UJ

3
f*\
«J
1

1
^
V)

~
1")
UJ
jmf
CD
H t














i

ft
49
e
!,
i ~ i
a
or 4>
• o
k C

— k
c
o
49
U






a
49
C
A
C

k
•K
g-
£


•-





t)
0
.c
49
I

°3?
0
I X
U
C

3
O*

k
U.
.

•
»*
li
It
>' 9









t
k
3
"S
U
O
k
OL
1


—
^
49
n
c
k
ITS
• —


*—
%» 9
• 3
49
T3 II
— -0
O 3

a

_C k
49 •
•• 49
B
*O k
• •
•D
> k
O •
k 49
a—

* U*
• •
k k
-1
— 3
U. C


kj

*> •».
•" t*
»- O
X •
u •
• k
0 0
o
Jt U
u
• ft
a: •
o
•

— •
••• c
• a
3 •• "
W ft
>2

c
o

49 ,
JI
•_
•
M
C
•^

k
^
turn
U.




















9
1
a
f
•
•Q
c
3

"g
•o
•
.?
*•
*

1
c
•
X

— • " •
51
H —
C 4)
I-H
0
• 49
j: »
n
~o c
o

k «
49 k

CL 49
^B
O S-






















































•
n
C9
C
"k

•D
•

k
3
C
Jf

C
0

JI
a
'•^














1

^
-C >
49 •

1^
k '"
O —
u c
o
k —
O 49
il
• |

Is
o o
W "D
C
o
•0 49
C li
• u
o
•
49 C
— O
C 49
•

9 »
49
• -
T9 O
C k
— •
li-
• •
(O *


9
"•" ^
.i-s
• j:
•i •
j: •
U 49
• •
• -o
JC •
0 —
££
u •


^ 0
— JZ

3

5 S



c c
o. o
49 *49

•O 49
~ *
0 g
> 3
U
• 0
1-^
• c
(O •

•
•

1
"
•
j:


•o
'o
»
•
J3
•
^
C "O
W 9
•o
• k
?0
u
•* A
49 k
• «
a o


• —
•o
- c
• o
• u
• —
k •
*2
O 3
— C
«•- 3








^
•
II
•
C
•
49
•


E
o
u

k
•2
































• k
• o
* •

•O k
•
— o
k u
• •
a. 49
•
u

49 -S
« C •
J~i
• O 49
• —
49 -O <«.
• ••
•O Jt •
k C
IS =






































u
•
k
k
o
u
y
^_
rt

•
•
1

2
o


a
E
•
•o
k
•
49
••
«•


Q

U

•^
•
O


9
m**
|

•
f
u
•

JC
u
•
u

X

^_
•
3
•
>








^
CL
E

«
49
•
O
a.
c
o

49
U
c
3
2T



•^
O
fl
U

^ '
49




C
o
49
u
c
3


m
g
k
•£
g
5
k
o



n
•
u

o.

a
c

•

1
K ^
II O)
k •
• -X

49 •

k
O k'
«•• o








c
o

*49
U
a,
n
c

*
o

^B •
49
«. u
o e
• k
• o
3 U
a
O-D
C
J p
e 1
k —
e -o
49 O
e k
o o.

^
• •-

49 *U
w «
,12-

49 X
a 49
k

!t o
0
— K
k
• c
CX.C
E 4>
OT *»




49
n
9
O

49
a

5
a
k
to
co

*• 5s
«*
JC ^
U *>
o c
51








JH
u
o

u
o

u.

•
u
o
I
•
ft

•
_c
49

5
•
k
ja

fl|
W












.
£
•
L.
O
«*-
















|
.






i














-------
                                                 Section  No.:  2.10.3
                                                 Date:  April 11,  1990
                                                 Page:  5
                                  Coarse Filter Holder
                                  Fine Filter Holder
                                     9.53 mm (3/8 in) Tubing


                                     6.35 mm (1/4 In) Tubing
Figure 3.1. Location of the filter holders on the sampling module.

-------
                                           Section  No.:   2.10.3
                                           Date:  April  11,  1990
                                           Page:  6
DIchotomous Sampler Field Data Sheet
Station Location t-AV'ti > ™ C. Run ^ate 12. [ |
Sampler Model 1# 4 £" S/N & ' 1
2 ^ SAROAD Number 332 3  b. ~£- 0
'•/ r. 0. W CFR /.4fc Umin
Vacuum Gauge Indications: Total Initial (^ Total Final I-!* Coarse Initial &
Coarse Final p

Rotameter Responses:
TSP* 13.*5 Final Total / 3 . 5" Ave. Total (T) ' '•
CSP' 'O.o? Final Coarse ^O.O Ave. Coarse (T) U
Elapsed Time Sampled /^^5" min
^ T5a ./f,i&^ , L/rni"
>.o Cr^i /• fr* Umin
TQa, CQa - 1 /m [T (Tav/Pav)1/z - b]
Total Act. Volume (TVa)- TQa X min sampled- ^ 5'*^ Umin FQa- /5". /^- i_/mjn
Coarse Act. Volume (CVa) « CQa X min sampled » 2
Comments: £}rfi.ss -n^c. i»^. ad(^a.c*^t -pf<
"* t^lOLOIK^ f^ i i,\fs * ^ "T™Y"aW\ ****\rn^_ ^^
•4/3 Umin FQa . TQa — CQa
/c( ; A. (oi" "»F Smofei,
^A*_pU^. o^-ite. o^-s-i-k.
•—*•""'
Operator CC^A. lAia.lv

Laboratory Calculations
Std. Volumes (Vstd): TVstd "2-3. S11) m3
TVstd, CVstd - (Va)CIO-3) [(Pav/Tav) (298/760)]
FVstd 2.1. 5X m3
FVstd « TVstd - CVstd
Filter Weights:
Fine: Gross Weight (Wg) loo . I3L mg
Tare Weight (Wt) '^.P-'l mg
Net Weight (Mf) 0- 92.5" ^
PM 10 Concentration ^ • ^ iiq/m3 nq/n
'Total or coarse set points, initial rotameter response.
CVstd "2 • 3T- m3
Coarse: Gross Weight (Wg) /OS. ^^mg
Tare Weight (Wt) Io4 . <4l 3 mg
Net Weight (Me) (?• ^S*/ mg
I3 -{Mf+Mc)(103)/TVstd
Figure 3.2. Example dlchotomous sampler field data sheet

-------
      3.
                                                            Section No.:  2.10.3
                                                            Date:  April 11, 1990
                                                            Page:  7
     Open  the  front  cover of the control module by turning the knob lat.'h
     counterclockwise.  The cover  is released by turning the "indicator one-
     quarter turn counterclockwise, and 'it is locked by reversina this
     process.                                                  • 3


4.   Switch the mechanical or digital timer to "ON."  If the sampler has a
     m,™ tal,*imer/Pr°9rammer equipped with a POWER switch,  turn on the vacuum
     pump.  Allow the pump to run for at least 5 min. to establish operating

            pp[e fCO"dltl°nS-  While th* Ampler equilibrates,  record on the
             !! Cation documentary information (i.e.,  location,  SAROAD desig-
             samp er model,  and S/N) and the run date of the sample   The

           ndVS 1Srat!Snhrelatio!ls!lips and the tota1  and coars* set  p°ints
     in Subsection 2         recorded.   Set point  calculations  are presented



5"   «!i  the nKta1 flow/ate by adjusting  the  rotameter'to the  calculated TSP
     value.  Observe and record the total  vacuum gauge  indication    The vacuum

                 fife? PreSSUre '^  (AP)  °f aPPr'°ximately 1  ?°  2  ]n   Hg  ?or
          value** SbservP^nd  "*" ^ adJustin9 the rotameter to the calculated CSP
          value.  Observe and.  record the coarse vacuum gauge indication- it should
          read approximately zero.  Turn off the sampler.    Incncai1on' u should
     7'   Ihl ™nPieV'S ri?W ^eady t0 Samp1e-  Set the master timer (according to
          the manufacturer's instructions  to energize the sampler for the next
          sampling period.  Reset the elapsed time indicator to zero?


     8.   Close the front cover of .the control module and visually inspect the

          SS'con^l51*; ^ensure that a11  ""-Pling components (sampling  nlet
          and control  module)  are in readiness for the next  run day.

3-3.2  Filter Recovery Procedure -



  '   lf   slmDler^lnn'  *T°rd ^ e]aPsed-time indicator  value  and  energize the
          record the Jn^ltnt^mplHr t0 Wam "P  t0 0Perati"9  temperature  and
          record the final  total and coarse rotameter readings  and  the final total
          a£ coarse vacuum gauge indications on  the  data  shief.   Turn the  sampler
           neall0n  Procedure and remove each filter one at a

          d shes    JeHfv  Jha?rf •??'   T?  1nhtheir Or1g1nal marked P1astic Petri  '
          dishes,   verify  that  filter  ID numbers match numbers recorded on the data



    3.    Calculate and  record  the total and coarse average rotameter readings as:



                              T - (TSP or CSP +'IF)/2                     (Eq. 10)

-------
                                                           Section No.:  2.10.3
                                                           Date:  April 11, 1990
                                                           Page:  8


where

              I - average total or coarse rotameter response, arbitrary units
       TSP, CSP s total or coarse rotameter set points, arbitrary units
             IF - indicated final total or coarse rotameter response, arbitrary
                  units.

     4.   Record the average ambient temperature [Tav  (K)] and barometric .pressure
          [Pav (mm Hg or kPa)] for the run day on the field data sheet.

     5.   Calculate and record the total and coarse average actual flow rates (TQa
          and CQa),  as determined by the sampler's calibration relationships.


                        TQa or CQa = l/m[T(Tav/Pav)l/2 - b]                 (Eq. 5)

where

     TQa, CQa ~ sampler total or coarse average flow rate, actual L/min

            I = average total or coarse rotameter response, arbitrary units
          Tav s average ambient temperature for the run day, K
          Pav = average ambient pressure for the run day, mm Hg or kPa
            m s slope of the dichotomous sampler total or coarse calibration
                relationship
            b - intercept of the dichotomous sampler total or coarse calibration
                relationship.

     Note:  Refer to Subsection 5 for a description of Tav and Pav measurements.

     6.   Calculate the actual fine flow rate by subtracting the calculated Qa
          coarse from the Qa total, and record.

     7.   Observe conditions around the monitoring site; note any activities that
          may affect filter particle loading (paving, mowing, fire) and record this
          information on the field data sheet.

3.3.3  .Sample Validation -

Validation Criteria - The following criteria have been established to assist the
operator in determining whether or not a sample is valid.

     1.   Timing:

          •    All samplers must turn ON and OFF within 1/2 h of midnight.

          •    All samplers must operate for at least 23 but not more than 25 h
               (1,330 to 1,500 min).                      '                 ~

-------
                                                            Section No.:  2.10.3
                                                            Date:  April 11, 1990
                                                            Page:  9
      2.   Flow Rates:
           The average flow rates must be within 7% of 16.7 L/min (total) and 1.67
           L/min (coarse) at actual conditions.  If these limits are exceeded
           investigate the cause.  Use the following criteria as a basis for sample
           invalidation:

                Decreases in flow rate during sampling (due to mechanical failure)
                of more than 7% from the initial  set point require a field calibra-
                tion check (Subsection 3.4).   If the sampler's calibration check
                indicates that the sampler flow was not Within ±7% of the designed
                flow,  the sampler should be invalidated.

                If the sampler flow rate decreases  because of heavy  particulate
                loading on the filter,  a post-sampling check  of the  vacuum gauges
                will  indicate  increased vacuum.   These filters should not be  invali-
                dated  because  they may  indicate  an  episodic  situation.

                Changes in flow-rate  calibration  of more  than  7%,  as  determined by  a
                field  calibration  check,  will  invalidate  all  samples  collected back
                to  the last  acceptable  flow-rate  check..  Recalibrate  the  sampler.

     3.   Filter Quality:   .

                Any filter that  is  obviously damaged . (i .e. , is  torn,  frayed, or has
                pin holes) should  be  invalidated.

3.3.4  Sample Handling -                                         •             •

Handling of a Valid Sample -


                              ' C°arse' and fine f1ow rates arrd complete the data
     2.   Promptly deliver the filter cassettes in their protective petri dish
          accompanied by the completed data sheet, to the analytical laboratory.

Handling of an Invalid Sample -

     1.   Complete as much of the data sheet as possible and explain any omissions'.

     2.   Mark "VOID" on the data sheet accompanying the filter and record in the
          site log book.

     3-   Do  not  discard the filter.
shee?^    '" f]e  Cassettes in their P^ri dish and the data
          shee                       K
          vanity  wniaS^adea1  lab°rat°r^  where  *  «""'  ^cision  on  sample

-------
                                                           Section No.:  2.10.3
                                                           Date:  April 11  1990
                                                           Page:  10


Handling of a Questionable Sample -

     If uncertain whether or not a sample should be voided:

     1.   Complete as much of the data sheet as possible and explain any factors
          that may affect the sample validity.

     2.   Put a question mark in the upper right corner of the data sheet.

     3.   Record as "Questionable" in the site log book.

     4.   Promptly deliver the filter cassettes"in their petri dish and the data
          sheet to the analytical laboratory, where a final decision on sample
          validity will be made.

3.4  Operator's QC Field-Calibration-Check Procedure        ,

     For dichotomous samplers, a field calibration check of the total  and coarse
flow rates is recommended after each month of operation.  The purpose of this check
is to track the sampler calibration stability.  Control charts presenting flow-
check data (indicated vs. observed) should be maintained.  These charts provide a
quick reference of instrument flow-rate drift patterns and will indicate when flow
limits (±7% variation from the indicated or design condition flow rate) have been
exceeded. - The"field check is made by installing a measuring device (which is
traceable to NIST and is calibrated within the .range of the total ,or coarse flow
rate) on the inlet of the sampler.  Calibration procedures for the measurement
device are referenced in Table 2.1.

     Calibration checks of the sampler flow rate require that the instrument be
running.  The following- flow-check procedures are specific to an orifice device.  A
variety of transfer standards may be used with this same procedure;  however, neces-
sary apparatus and subsequent calculations to determine the sampler's flow rates
will vary.

3.4.1  Ffeld-Check Apparatus -

     The following equipment is required for a field calibration check:

   • •    A thermometer capable of accurately measuring temperature to. the nearest
          ±1°C and referenced to an NIST or ASTM thermometer within ±2°C at least
          annually.

     •    A barometer capable of accurately measuring ambient barometric pressure
          to the nearest ±1 mm Hg and referenced to an NIST or ASTM barometer
          within ±5 mm Hg at least annually.

     •    Two calibrated orifice devices and calibration relationships (one for
          total  and one for coarse).

     •    The sampler's calibration information..

-------
                                                           Section No.:  2.10.3
                                                           Date:  April 11, 1990
                                                           Page:  11


          Two clean flow-check filters.

          Dichotomous sampler flow-check data sheet (Figure 3.3) or log book.

3-4.2  Procedure for Field-Calibration Check -

     1.    Insert clean filters (designated  "flow-check filters") into both  the fine
          and coarse filter holders  of the  sampler as  described  in the operating
          procedure in Subsection  3.3.1.  Flow-check filters  should never be used
          for subsequent  sampling, as  particles  larger than 10 urn can impact on  the
          filter when the inlet  is removed  and bias the sample.

     2.    Turn  on the sampler and  allow it  to warm up  to operating temperature
          (approximately  5  min).

     3.    Read  and  record the following  parameters on  the sampler flow check data-
          sheet  (Figure 3.3):

              Ambient temperature (Ta), °C  and  K

              Ambient barometric pressure  (Pa),  mm Hg  or kPa

              Sampler S/N  and model                          •               .

              Orifice S/Ns and calibration  relationships

     -    •    Date,  location, and operator's signature

              rrop1rc rotameter's calculated flow rates  and set  points:  TFR,  TSP-
              CFR, CSP.                                                          '

    4.   Adjust both the total and coarse rotameters  to their respective
         calculated set points (TSP, CSP).

    5.   Remove the inlet from the sampler, replace it with the  flow-check  orifice
         device, and recheck the rotameter  set points.                     unnce


    6'   2eflec?ione^?2?haCh°t%S th%tojal  flow  orifi« ^ reading the manometer
         inn fin    
-------
                                                             Section No.:  2.10.3
                                                             Date:   April  11,  1990
                                                             Page:   12
                     Dlchotomous Sampler Flow Check Data Sheet
Station Location.

Sampler Model .

.Pa 14 "Z-   mm Hg       Ta
               Date.
                 S/N
                                                     SAROAD Number 17 2^01 Eft

                                                       EPA Number
                             K
                                                       Unusual Conditions
.Orifice S/N  ll"7
     Orifice S/N   "Si b (t»+ne^ Orifice Calibration Date.
Orifice Qa (Total) Calibration Relationship:

Orifice Qa (Coarse) Calibration Relationship:

Sampler Total Calibration Relationship:
                 ,  TFR
Sampler Coarse Calibration Relationship:
                                                  L/min,
                               L/min,
                                                             TSP
                                                             CSP.
     Flow
     Rate
'  Description
AH.O
 (In.)
                                Orifice Flow
                                   Rate*
                                  (L/mln)
 Sampler"
TQa or CQa
  (L/mln)
                                                 Difference
L/mln
Total
Flow
        Design
        Cond.'
                                16.7
Coarse
Flow
        Design
        Cond."
                                1.67
* TQa or CQa « m[(AP) (Ta/Pa)]1/2 + b
."TQa or CQa » 1/m [(TSI? or CSP) (Ta/Pa)"2 - b]
                QC % Difference - (TQa or CQa) - Orifice Flow Rate ' (1 00)
   -Design condition % Difference
Operator
                      Orifice Flow Rate

                    Row Rate - (16.7 or 1.67)  (100) '
                         (16.7 or 1.67)
            Figure 3.3.  Example dlchotomous sampler flow-check data sheet.

-------
                                                             Section  No,:   2.10.3
                                                             Date:  April  11,  1990
                                                             Page:  13


      8.   Observe the AH20  across the coarse orifice by  reading the manometer
           deflection and determine the corresponding flow  rate from the  orifice
           calibration data.  Record both values on the flow-check data sheet
           Using the sampler's calibration relationship,  calculate indicated coarse
           actual flow rate  (CQa) and record.

      9.   Using the above information, calculate the QC  percentage difference as:

           % Difference = (TQa or CQa) - Orifice flow rate   Mnn.            ,_     .
                                  Orifice flow rate          (-L(J(J)            (Eq.  11)

      10.  Determine the percentage difference between the sampler design  flow rate
           (16.7 L/min or 1.67 L/min)  and the orifice determined flow rate as:

                        = Office flow rate - (16.7 or l.fi?)  ,.n_.          .    '
                                 i   16.7  or 1.67  	~  (10°).         (Eq-  12)
                             *       3s  within 93  to 107% of'the 16.7  L/min  or 1.67
                      T"? th(at  a-tUal conditions).  the sampler is  operating
           ™l,n         ?5eSe   imUS 3re exceeded'  investigate and correct any
           resumeS           necessary'  ^calibrate  the sampler before  sampling is


      12.   Turn  off the  sampler,  remove  the  orifice  device,  replace the  inlet  and
           reconnect  the fine flow  vacuum line.       .                        '

      13.   Remove the filters from  both  fine  and coarse  filter  holders.
3.5  Documentation
3-5-1  Operator Who Starts the Sample -

     Mark on the filter petri dish:

     1.    Sampler ID number.

     2.    Fi 1 ter number.

     3.    Sample date.

     4.    Designation  [e.g.,  whether  it  is  a  coarse  (C)  or  fine  (F)  filter].

     Mark  on  the field  data sheet  and  record  in the  log  book:

-------
                                                           Section No.:  2.10.3
                                                           Date:  April 11; 1990
                                                           Page:  14

     1.   Site designation and location.
     2.   Sampler ID number.
     3.   Filter ID number.
     4.   Sample date.
     5.   Initial flow rates and rotameter readings.  Initial temperature and
          barometric pressure, if required.
     6.   Unusual conditions that may affect the results (e.g., subjective evalua
          tion of pollution that day, construction activity, weather conditions).
     7.   Signature.
3.5.2  Operator Who Removes the Samples -
     Mark on the field data sheet and record in the log book:
     1.   Elapsed time of the sample run.
     2.   Final flow rates and rotameter readings.  Final temperature and baro-
          jietric pressure, if required.
     3.   Existing conditions that may affect the results.
     4.   Explanations for voided or questionable samples.
     5.   Signature.

-------
                                                            Section No.:  2.10.4
                                                            Date:  April 11, 1990
                                                            Page:  1
 2.10.4  FILTER PREPARATION AND ANALYSIS
           ^n •  y*Z  a ™i;.sa?P '"S  Program  depends  on  several factors.  A primary
 th™    ?h   the analytical  laboratory  staff's  attention to detail and balance
 coercion', J iL?eCi;°n °fferS  Sidelines  to enhance the accuracy of laboratory
 operation and hence the mass  concentration  determinations of PM10L
                            i's1on' calibration requirements, and recommended  filter
                   f
 4.1   Filter Handlin
4.2  Filter Integrity  Check
     Specific defects to look for are:


                               apPeTng (a)  as a ^stinct and  obvious bright
    2                                          lack  of  seal  between the fmer and
    3-     aff °r nashinq--An.y extra  attached  residual material on the reinforc
                           -                                            "

-------
                                                           Section No.:   2.10.4
                                                           Date:   April  11,  1990
                                                           Page:   2
50-mm Diameter
Plastic Petri Dish
with Tight-Fitting
Lid
37-mrnDlameter
TeflorFFilter
with Polyolefln Ring,
in Cassette
          Rgure 4.1.  Dlchotomous filter cassette and petri dish.

-------
                                                       Section  No.:  2.10.4
                                                       Date:  April 11,  1990
                                                       Page:  3
      4.   Loose material-Any extra loose material or dirt particles on the filter
          that require removal by brushing prior to weighing.             "ner


      5'   ofSa0contaam?nan^y °bV1'°US VlSible dl'scoloration that might be evidence


     '6.   Filter nonuniforaity-.Any obvious visible nonunifomity in the appearance
          of the filter when viewed over a light table or black surface that mioht
          indicate gradations in porosity across the face of  the filter.       9

      7.   Other-A filter with any  imperfection  not described above  such as irren
          UTIr-surfaces or other results of poor workmanship.                   9"

 4.3   Filter  Equilibration

      Filters must be  equilibrated in a conditioning environment for at  least ?&.  h

 valuerebetwee9nW|d9,heHd-^?ela^hVe  """^I* (RH) shoufd bi  KlTJSnstln? aJ fLw
 chnn?H h   t U     S  45-s, with a variability of not more than ±5%.   Temoerature
 o? not more' haS°*n3Sorant ?lth.a mea"^lue between 15 and 30'C, with  r^riaMmy











t®2^&r~Z32S^&&s


4-4   Initial  Weighing  Procedures  (Tare Weight)
    NOTE:  Make sure that the balance has been calibrated (at least .ni.u.11,,1
          ma,nta,ned according to manufacturer's recoraendations   I? out of

               "™  HaV                       "cording tS

-------
                                                            Section No.:   2.10.4
                                                            Date:   April  11,  1990
                                                            Page:   4


      2.    Zero  the balance according  to manufacturer's  directions.

      3.    Have  the QC supervisor  perform the "standard" filter QC check  (Subsection
           4.5.2)  to  increase the  validity of subsequent tare weight values.

      4.    If  filters must be weighed  outside the  conditioning chamber, use caution
           to  avoid interference with  ambient hydroscopic particles .and begin  the
           weighing procedure within 30  s.   Weigh  the  filter according to  manufac-
           turer's directions, making  sure a stable  reading  is obtained.

      5.    Place the tared filter, with  the reinforcing  ring side  up,  in a
           comparably sized petri  dish.

      6.    It  is recommended that  each balance be  assigned a block of filter numbers
           to  be used sequentially.  Assign a filter ID  number and take extreme  care
           to  avoid duplication or missed  numbers.

      7.    Legibly record the assigned filter number on  the  petri  dish, leaving
           sufficient room for one more  letter designating size  fraction [(F)  for
           fine  or (C) for coarse] to be written following the number.

           Suggestion;  The operator may decide to include three additional digits
           on  the petri dish label to represent the  tare  weight  of the filter  (e.g.,
           101,  99, 105,  etc.).                                                  -

      8.    Record the balance number, the  assigned filter number,  and the  tare
           weight on the data/coding form  (Figure 4.2).   Number each form  sequen-
           tially in the upper right-hand corner.

      9.    Perform regular QC checks as  detailed in  Subsection 4.5.

      10.   Install  the filter into a cassette, and return the  filter/cassette
           assembly to its. individual petri dish.

4.5   Internal QC

4.5.1  Operator QC -

      After every fifth weighing,  the operator should recheck the zero and calibra-
tion  of the balance and  record these check values on the Laboratory Internal
Quality Control  Log (Figure 4.3).   (The zero and 10-mg weight checks are  internal
standards  of the analytical  balance.)    Zero QC checks  within 4 /*g of true zero and
calibration QC checks within  2 fig of 10 mg are acceptable.  Larger discrepancies
should be  corrected immediately.   When QC checks are unacceptable, the previous
five  filters must  be reweighed.  Any filter weight outside of the normal  range of
80 to 110 mg must  be investigated  immediately.

      Note;  An electrostatic  charge will prevent a microbalance from operating
properly.  Static  charge is  the accumulation of electrical charges on the  surface

-------
Section No.:
Date:  April
Page:  5
 2.10.4
11,  1990
                   CD
                   5
                   o
                   i
                   fr
                   o
                   5
                   &
                   a
                   x
                   u
                   CJ
                   21

-------
                                                         Section No.:  2.10.4

                                                         Date:   April 11,  1990

                                                         Page:   6
    o
    o

     5

     E
i
=
i

-------
                                                             Section No.:   2.10.4
                                                             Date:   April  11,  1990
                                                             Page:   7


  of  a non-conductive material.  Common symptoms of this  problem  include noisy read-
  out, drift, and sudden read-out shifts.  To reduce static  charge within  the
  ofn?""'^1 may- !?e necessary *° Place a radioactive  ionizing  unit  (i.e.,  Polonium
  210) in the weighing chamber.  It may also be necessary to pass the filters over an
  ionizing unit before they are weighed.  For more information  about  static and how
  to  minimize its effects, see the.Technical Note, "Static Control for Balances "
  prepared by Cahn Instruments, Inc.

  4.5.2  Supervisory QC Procedures -             .

      1.   Keep a bound QC notebook.  These notebooks must  contain all  QC data
           including the balance calibration and maintenance information,  internal
           routine QC checks, and independent audits.   It is recommended that con-
           trol charts be maintained on each balance and included in this  notebook.
           These charts may indicate any excess drift  that could flag an instrument
           malfunction.

                                         all  QC data  on the Quality  Control
           fhp1™®91""1'!!9"-0?-^—1'9!!1'"?  day'  after the 0Perator  has  completed
           the  zeroing  and  calibration  checks of the  balances,  tare weigh one
           arbitrarily  selected  filter  from a set  of  "standard"  filters  (10% of the
           total number of  filters to be weighed).  Because these  f Uteri represent
           LSB     J QC  check  do not use  them  for subsequent sampling.  Thlse
           weights must be  repeatable for each balance to within 20 ug of the
           original value.   If not, the balance performance is  unacceptable-
           troubleshoot and  reweigh the filters as necessary.   If  more than one

           Jhi?n!£tI™-neSVhake  -a-e ?hat the (iUer  1s we^hed on the same
           that determined the original tare  value.  Unless  this procedure is
           adhered to,  many  samples may have  to be invalidated.

           Reweigh five to seven exposed and  unexposed filters per balance each dav
           of operation   Weights should be within *20 /.g of origina? values? if
          .not,  troubleshoot and reweigh.   Because of the loss of volatile

                           ''
           hv   ?° I1'1"]'1? a1 -Set for exP°sed filters.  Record all data on
          the Quality Control Log Form and in the QC notebook.
     5"   th^ilL3^6^1']1^ 2t fl'iter wei9hts and data completeness daily on
          the laboratory data/coding forms and initial.  When bound  these serve  as
          a laboratory notebook.  Sign each completed form.               -

4-5  Post-Sampling Documentation and Inspection


this p?ocedr!re:'Pt  °f ^ Sample ^ the field<  the Samp1e c^todian  should  follow


     1.   Examine  the field  data sheet.   Determine  whether all  data needed to  ver-

-------
                                                           Section No.:  2.10.4
                                                           Date:  April 11, 1990
                                                           Page:  8


     2.   If the exposed filter was packaged for shipment, remove the filter from
          its protective pertri dish and examine the petri dish.  If sample mate-
          rial has been dislodged from a filter, recover as much as possible by
          brushing it from the petri dish onto the deposit on the filter with a
          soft camel's-hair brush.

     3.   Match the filter ID number with the correct laboratory data/coding form
          on which the original bal-ance ID number, filter ID number,  filter tare
          weight, and other information are inscribed.  The sample custodian should
          group filters according to their .recorded.balance ID numbers.  Initial
          separation of filters by balance ID number will decrease the probability
          of a balance error that could result from the use of different balances
          for tare and gross weights.

     4.   Remove the .filter from both the petri dish and the filter cassette.  The
          filters must be handled with clean, nonserrated forceps; they must not be
          touched by the hands.  Inspect the filters for any damage that may have
          occurred during sampling.  Reject the filter for mass concentration
          determination or any additional analysis if defects are found.

     5.   Return filters with no defects to their original petri dish and forward
          to the laboratory.  File the data sheets for subsequent mass concentra-
          tion calculations.

     6.   Return defective filters with the type of defect (or combination of de-
          fects) to their original petri dish, labeled by defect type(s),  and
          submit to laboratory supervisor for final approval  of filter validity.

4.7  Final Weighing Procedure (Gross Weight)

     1.   Group filters according to their recorded balance numbers.   (Filters
          should be separated initially by balance ID number;  this will lower the
          incidence of balance error that would occur if different balances were
          used for tare arid gross weights.)  Reweigh each filter on the same bal-
          ance on which its tare weight was obtained.

     2. .  In an environmentally controlled area, open the petri dish, making cer-
          tain that the lid (with the filter ID inscribed) is  placed  beneath the
          bottom and that no mix-up occurs.

     3.   Cover the open petri dish with a clean laboratory paper towel and place
          it in the conditioning environment.  Allow the filter to equilibrate
          according to procedures outlined in Subsection 4.3.

     4.   Repeat Steps 1 through 5 of the dichotomous filter tare weighing proce-
          dure (Subsection 4.4).

     5.   Perform the  internal QC checks described in Subsection 4.5  to ensure
          validity of  reweighing.

     6.   Record the indicated gross weight on the laboratory  data/coding  form.

-------
                                                            Section  No.:  2.10.4
                                                            Date:  April  11, 1990
                                                            Page:  9


     7.    If the dichotomous filter is not to receive additional analysis, place it
           back into the corresponding petri dish.  Deliver  weighed  filters to the
           sample custodian for archiving.

     8.    If the filter is to receive further analysis, return it to the petri dish
           and note on the petri dish what additional analyses are required.  Place
           an asterisk after the gross weight column on the  laboratory data/coding
           form to indicate that the filter requires additional analysis.  Carefully
           place each filter thus packaged in a box, and deliver to the sample
           custodian who will forward it to the laboratory responsible for the
           additional analysis.

4.8  Calculation of Net Mass Filter Loading

     The gross weight minus the tare weight of a dichotomous filter is the net mass
of the particulate for that filter.  Each calculation of this process must be  inde-
pendently validated.  Refer to Subsection 5 for information regarding the calcula-
tion of PM10 mass concentration.

-------
Section No.:
Date:  April
Page:  10
 2.10.4
11,  1990















Wt
I
Ci

t/)
IHI
tn

<
5
JJJ


i
M

Jj
£C
g
£
cc.
Ul
J"—
_j

LU


•
*F

Ul
—I I
CD

}•*•



























•*>
i
_!_%
p
— •
c •
0
u c

u

1
u


•



n
4}
c

*
u
3
cr

cc













>•
c
•
3
er

t.


T)
C
•

•o
o
X
^
1
1
1
T







*>
"~

A)
0
<







^
3
er

c
1

^
o
u
1


«
•0 I.
C • N
•££
• — m
a.*- —
• T)
u •
L U —
out.
». — 4J
0. •
T> Q.
• • >
U • —
1. > *>
• O O
• — *
0 O
ecu
o a

li.s



•

u
3
5
o

a

a
c

•H
•o
c

X

0
L.
2

o









a
C
•»
•MI
^O

X

















X

•KM
N
M»
jj
C
3
0"

t.
* -

g
c

"O
e

















































d
.8

^
^
u
n
0



u
•a
0

«
n
t.
•*
K

<••*
O
X
e

"a.

1

.
U
tf

-M
•*_
X
u

0

4)
u

a,

c


X
^B
•J
3

>



.X
U
•
X
u

X

t_
01
•

c





t
_x
T
•
o.
o
1.
a.
X

c^
^J
1
M
•O
* e
— o
x- •*> e
•- • o
O L.
• O -D
— J3 C

X ~ E
U 5 °
1 c at •
OCX
n — •
S -a -o —
• O T)
u t> o
• L •**.—
4J O • <-

UL i. -D a.




L.
f
^j
~

X
u

0 •

^K 0
u c
• a

u *

X •

— 1.
3 JO
• t

> C




c
o


•
u

IK
4J
c

T)




C
o
t.
X
*3
•

^J
^
|
ce


— ce
o
L. •-
•U JC
C
o -c
U M
*.
O
c •»•
o

-*> c

i 1
•«• O
*H* ^
•«• *^
3 >
er c
Ul •

t i
— c
— o
3 U
er
• t.
11

JS
•o o
L.
o, u
u o

* E
o
•e o
C L

c
• 0
u *>
• u
X X
o —



«
M
C O
0 1.
-*• a

• -o
u c

•^
« 0
•— u
3 •
cr*»
Uj «_*














e
• wo
• 10 —
• E S *>
— o m m
JC -O U
*> TJ c a
— c • >

1A O
K e --o
10 o o>
•Q • • C
C — * <•
• L. -t> X
• • *>

CM •
K • «
C LO U •
• +1 3 —

3t e • x
*>•!.*»
• X • —
X *> & >

E X
3 U
E •

c
— u
E 0

-fc> O

• 'l.

<- o.

« c
X O
o —

£
• L.
M J& •
§ = .2
-•sf
— o- •
•o • n







c
0


u
u


•o
1
1.
JO
'£
"3
IT
1
u

1
X
CO

1-2
ce *>

•
•5.
*> 7iii«> •
X X •
oijj e •» L
— M — •
• • X E X
k t. *» o- E

-u a t»
— o c e e
0- X > —
TJ a o 4)
•o * — § •
9 C • * L.
•«» — * 1. -0
• E
o u c t. —
•«• ^ »•• ^ •«•
-O -iJ CI-J 3
c • e •» er
H-I -o aj <• •
•


••

c
3
• CT
U
O —
L •
a c
L.
9 •
C •»

X —
a


r • •
t_
0 E 3
> C "O
<- o •
• «- u
« t. O

o a. a.





a
c

X
a •
— u
e 3
* "O,

• u
u o
• i.
t- a.

n
9
t .
• i.
• O
M
IE
o •
<«- 3
. 1- II
'.2
X
44 -P
U
T> O
— a
o •
> c.


•« e» c
' e o
u
*> — e .
• CO
o —

• '*> u
LUC
• •»» 3
s s •*•
-:Il
• u
• o %• •

o o
JC • • —
*> o
e • c •

"a"aT3 "a.

i? s'5 :
-o
c



0
^
1
«

_•

^ .

«M( •
U C
• o
X —
u *>

X ^
— c
-; ?
3 3
n u
— o
> -o

i
1 3
U
O TJ
a-o c
c • c
.— • o
— e -•-
to c *>
— 0 •

II O -»J —
1 • • «-

• net
o c « e
o. — i >


i
L
•^
^
3
•

|
X
O9

jj
ce 4>
a

•o »—
• — •
c • >

°* * i

X *— 1.
o a
-li
•&..S
•54*
* •

"O ^1 02
• m
*) 49
* n e
u • —
— L. X
•O • *>
C * —
>-« c »
1

0 O1
o

a •
c
O) 1.
c c
X C

..

* —

L

^ U
— O •

 ^
(-•-.•
• c u
« u o
X 3 L-
O -O Q.

•


Ol
c


O)
e L.

«
•• •
« u
o o
1. U
O *Y
















t.
"i
•
X
u

e
o
*>
•
u
JO


.^
3
er
•

E
O
b
«»•











































-------
                                                             Section  No.:   2.10.5
                                                             Date:  April  11,  1990
                                                             Page:  1
  2.10.5   CALCULATIONS,  VALIDATIONS,  AND  REPORTING  OF  PM10  DATA

      Measurements of  PM10 mass  concentration  in the  atmosphere that  are  used to
  determine attainment  of  the  National Ambient  Air  Quality  Standards for particulate
  matter must be expressed in  units of micrograms per  standard cubic meter (ag/std
  mJJ of air.  For these measurements, "standard" means  EPA-standard conditions of
  temperature and pressure, which are 25  °C  (298 K) and  760 mm Hg  (101 kPa)  respec-
   K- y:  J?IS sectlon Presents  the calculations required  to compute  and  report
  ambient  PM10 concentrations.  A summary of all calculation formulas  and  associated
  symbols  presented in  Section 2.10 is given in Table  5.1.

      Particle size discrimination by inertia! separation  requires that specific air
  velocities be maintained in the sampler's air inlet  system.  These design veloc-
  ities are obtained when a specified "design flow  rate" is maintained.  The design
  flow rate is specified as an actual  flow rate (TQa and CQa) ,  measured at existina
  conditions of temperature (Ta) and pressure (Pa).              *"urea ai existing

      The sampler's operational flow rate (i.e.,  the actual flow rate when the
  sampler is operating normally to collect a PM10 sample) should,  of course  be very

        °   '110"     '      ^ ^?]er3  hav? Some means for
o erat on   now             ' *          ^?er3  av? Some means for ^asuring the
       ?!? „  ?? h f '  J"J-t5a3,flow rate measurement system must be calibrated .
      ically with a certified flow rate transfer standard.   Usually  measurements
                                                          .                       s
      nt.           temperature and b*™metric pressure are requred to gel
      f  n   indication of the operational flow rate. . For determining the average
      er flow rate .over a sample period,  use of average temperature (Tav)  and  a-ver-
  Pmnprr?m^rlCHPreSSUre (Pa^  °vr the Samp1e Period is recommended    If  averagf
 elch  samoH npHSH65^"6 Va1U6S (°P !Teasonab1e estimates)  cannot  be .obtained  for
 D?essureP  PO  fnr ;H.    rr>  ^^l 3Verage temPerature (Ts)  and barometric
 pressure (Ps)  for the  site may be substituted.

 fmm  ?V *"%  ?f,\ reftd " '"9S  ?**  be recorded on-site  or estimated  from data  obtained
 station   £?  U^--Natlona1 Weather Service Forecast Office  or  airport weather
 be  It  J;at?Sn  n     C pr?SSUre  readin9S obtained  from airports or  other sources must
                    Ur
                                                                               u
                f       (1'!-'  "Ot  corrected  to- ^a  level),  and they may  have to b
                             **"
corrtP   n     f                                         ,
theliroort   ?} {ndT^  ¥?**" 5"? elevat-ion  of th* monitoring siteand that of
period and sea^nnll II »     f  "jj  P'V  readin9s cann°t be obtained for each sample
?aken that ?K Srf  ?vjr 96S  .   the,S,!te are  routinely substituted, care must be
taken that the actual temperature and barometric pressure at the site can be









may £ substituted  for Tav, and Ps may be substituted for^ ?n Equations l]  I,

-------
Section No.:
Date:  April
Page:  2
 2.10.5
11, 1990














S
I
e
0

.*>
0
e

j&
3
to
C
0
3
1
f
1
i ^!
^
a !
0.
z

£

o
IS
•<

O
§
1^

W
**C

5
«•*
0
u.

«4
*

UJ
*"


i
!

1
i
i





^4

CM




"
er
Ul



I
I
1
1
f







•
3

k
0
u.

















Calculation

**>





•J
t-

n '
r*

TS

fl
Q.
^«»
CL

cr
u


cr



•

•^
k
Conversion of
f low rates from
actual to atanda
conditions



•

vH

CM



ft
•"
er
UJ









X^

f
"^,

m
t~
^^

•0
jj
ii
o.
^^
>
a.

ley
u


icr

_•
e a
a 3
m Ji
k U
e a
Conversion of av
flow rates from












































n
c
o
• «•
^
•••>
to standard cond



^ ^

•-I CM CM

CM CM CM




O4 <*J ^C
• - • •
er er -er
UJ Ul HI






JO

*
^•^

*> k
A e
>- -o
•». c
•D e
M •

>-* k
? CM £
O. ~». M
•>, .•» c
• *™^ e
1— • k
^ o. •*>

•o e •
« t 2 "
cr t-i e

u u H
• o o
cr •< <

. c
c - o
n o «-
-e c — > «.
* k O *>
o • •- e c TJ M
— -o •> e— koi-c
«-C— • • J — • O
• -a k 3 — jj -o —
H. *> C • ^> — u C *>
0 • 0 O O •«• —
o o • • o u *> -o
c E c k • c
00— TJ k O kO O
— k • ce— euku
•lv-3 vl^^3 «> e
k *> eu e — «.—
e*u — E e 6 « « «
> • • ««k «3C3
c^> ^«>k •& +> m 4)
o m o ooo ouko
 (— k o cea^a



v ^ f ^r ^

M CM CM CM CM

M CM CM CM CM




U) « r> CO 0>
er er er er er
Ul UJ UJ UJ ill






JO
1
CM
** CM CM

> M • *H
a. ^^ T I?
•*•• • tl H- 1—
> t- »- ^. ^
t H H ' fc t


•— • 1"i" ** ^
_E: ' a. o. * *

^il MM ^^ ^^
|| tt * K
H. O

ff « * £. >£
u «" •* "T*
O It II II ||

« ce ce a. a.
er u. u. co co
l- . t- o t- u





i e

*)' «» e k « c k
c. o n .a n ko C
0k— k e a *>
u x • — • Z ti a
L. *> o • o ee £
• —00 U Ik •
*> A. « • *>
c • -o e -o e *> *> o
— ec*> c*j oc k
c e • • <• k —
o-Z — u u o e
a. « * c ek • k
a'V'tfoo -^o ^i'4> a
— co — — o— oe o
coa*>»--*> )-*. i- • u
















































A
yi
c
o
a.
«
«
k
*9
C
'o
a.
4*
o
»



CM

€«>

m



IB

UJ














CM
U?
hH


ft.
CO

k
o
a.
CO
%_^

II






f
Total and coarse
average rotametei
readings



CN

^

en



^4
•H













er
UJ


f
^.
e
k
• *
j>

e
u

^
k
o
1
4|
cr
•



i

L
*
Q


C
U
«*.
•••

oio .
1
k
o

a




a


n
•*
H-




o









^

3
C

jj
C
o
u






































!

i
QC percentage
difference


-------
Section No.:   2.10.5
Date:  April  11, 1990
Page:  3














*•*
3
C
^
C
O
u



.
u>

UJ
_j
0)
LK








1











c
! .2
*>
u
•
1 «
3
C
O
t •—
1 a
i O"
; iu


i
i
i
i --









- 3

^
O
U.











. ,
i.
1
i
1

C
O
*J
^
u
•
'
"
W *•< • C4 C« -C T

*f *"*»•«** ^ «f * *
• * • • , ' ^
<*5 U3 U) l£ ID l£ 9"- f»>»
2 2 2 "> "o "• °° «
*^. f-l«-« |^ ^« g^, , ,_|

^" cro'tr^trcr cr
"J UJlUUJUJuJuj UJ
§ . i
** ^
? 1 ' i !
I 1 • i '•••'!!

|s'-. ^ ' 7 • ^ *' '! '
*° ^ s' ' >^ mi m' ' '
C. * O '"I •**! Oi -3 i ;
^ 1 . •*•> av< • !
S^: q - -i •• • ;
*'° ' v- iiL T? ii_ »*
"^ -o^ ^'i •»*• "S-*" "••«••
CO ^ ' *' !— «
*'« • ^T1 *l 01 IT) o c
*:^ 53 J !ii 1= j =« |.
^ 1 f- » 5 ^^ ^^ '

i iit' *^ *~* *^ 9'. u -— ^ *
9 it ^' j"' — iQf *3' Q i
U II ^" il * §! ^5 '
•H' -^ ' 5 c ; •*> i JO' «i . ri
•~ jJ • O C*> > ^ .1 ' *••* t^j .T! I
o a '"* *'Z B' - m' «' •'
S T? — • C. B; ** ar-
11 *> • C «-: cy Ci „ „
>- « •»>••> 2- UL ul 11 . II
• o ar o c • *-* 2,
*" '!-<»• "" „. '
.t 3 r * g » « - t
Q i" n • 2 S ^ „ s S
v, ^ 3 O U. 0 °
**. 'K > o. >_i k_, . J* x
— « ' !
« • c , :
"- "D 00 3 O .' , i
il*C iS3 ' • .i.-S.J,''!:
•2 S5.f ssi- ^ -5 -5 • ii-^ii-S-2 |
»-««6 ooo ° "z i2*l-i:2ii'-
•--•-i 'T'-0 . e • =• co •- -o • ± J :I
IfiJ. s:s'J- :'l 11 il I?1^ 'Isl 1!
'ij'cS^ '" • "= > i" S^ i^o i°~§ ^"SJ"? J-I^J. •

2sl=-s is- 3: •§ ' l-ii Hi 5t«i li.?-s

-------
                                                            Section No.:  2.10.5
                                                            Date:  April  11, 1990
                                                            Page:  4


 5.1  Calculations

 5.1.1  Flow-Rate Calculations -

      The total and coarse flow rates are calculated by first averaging the
 sampler's initial rotameter set points (TSP or CSP) and final indicated rotameter
 responses (IF).
                               I = (TSP or CSP + IF)/2
                                                                     (Eq.  10)
 where
               I s average total  or coarse rotameter response,  arbitrary units
        TSP,  CSP. = total  or coarse rotameter set points,  arbitrary units
              IF * indicated final total.or coarse rotameter response,  arbitrary
                   units.  .                                                     J

 These values are then  applied to the sampler's  total  or  coarse calibration rela-
 tionship.
                         TQa  or CQa  =  l/m[I(Tav/Pav)l/2  .
                                                                     (Eq/5)
where
TQa, CQa = sampler total or coarse average flow rate, actual L/min
       I - average total or coarse rotameter response, arbitrary units
     Tav - average ambient temperature for the run day,  K
     Pav = average ambient pressure for the run day,  mm Hg or kPa
           slope of the total  or coarse flow-rate calibration relationship
           intercept of the total  or coarse flow-rate calibration relation-
           ship.
              m
              b
     The average flow rates are then corrected to EPA reference standard conditions
by using Equation 13.  Note:  For the subsequent calculation of PM10 concentra-
tions, only the TQstd flow rate is considered.

             TQstd or CQstd = [(TQa or CQa).(lQ-3)(Pav/Tav)(fstd/Pstd)]      (Eq. 13)
where
     TQstd or CQstd =

         TQa or CQa =
               10-3 s
           Pav, Tav =
         Tstd,  Pstd =
               total  or  coarse  flow  rate  corrected  to  standard  conditions
               std. iTH/min

               s.ampler total or coarse  flow  rate, actual  L/min
               conversion factor for  L/min to m3/min
               ambient barometric pressure,  mm Hg,  kPa; temperature   K
               standard  temperature and pressure 25°C, 298  K, 760  mm Hg, or
               1U1 kPd•

-------
                                                            Section No.:
                                                            Date:  April
                                                            Page:  5
                                                            2.10.5
                                                            11,  1990
 5.1.2  PM10 Concentrations Calculation -

      The reporting of total PM10 mass concentration data requires the calculation
 of the total  volume of air sampled (Equation 14)  and the computation for total  mass
 concentration (Equation 15).
                                    V = (TQstd)t
                                                              (Eq. 14)
 where
          V  =
      TQstd  =

          t  =


      PM10   =
total sample volume in standard volume units, m3
total flow rate corrected to standard conditions (see Equation 13)
std. nwmin                                                       '
elapsed total sampling time, minutes.
  (Mf + Mc)(lQ3)
       TV)
                                                                            (Eq.  15)
where
      PM10 = mass concentration of  PM10, /
-------
                                                            Section No.:  2.10.5
                                                            Date:  April 11, 1990
                                                            Page:  6


      2.   Compute the total mass concentration of PM10 for seven samples per 100
           (minimum of four per lot) as specified in Subsection 5. Y.I or 5.1.2.
           These suggested starting frequencies may be altered, based on experience
           and data quality.  Decrease the frequency if past experience indicates
           that data are of good quality, or increase it if data are of poor qual-
           ity.  It is more important to be sure that the validation check is repre-
           sentative of the various conditions that may influence data quality than
           to adhere to a fixed frequency.  If calculation errors are found  all
           values in that sample lot should be recalculated.

      3.   Scan all total mass concentration values,  note those that appear exces-
           sively high or low, and investigate.  Repeat Step 2 for these samples.
           Compare validated PM10 concentration to the originally reported  value.
           Correct any errors that are found,  initial  them,  and indicate the date  of
           correction.

      4.   If all  mass  concentration computations  appear correct  and  questionably
           high or low value(s)  still  exist,  review all  raw  data  (i.e.,  sample time
           average actual  total  flow rate and 'its  subsequent correction  to  standard'
           conditions,  and total  net particle  mass for coarse  and  fine filters) for
           completeness  and correctness.

 5.3   Data  Reporting

      The primary  standards for  paniculate  matter in  the  ambient  air are based on
 the measured mass  concentration of -PM10. '  Information  on  reporting and  interprets-'
 Jrt°?rS  BJ  9 data  with  respect  to the-attainment  of these standards  is  covered in
 40 CFR  50, Appendix K.

 5.4   Additional Calculations

      This  section  outlines the procedures and  computations  necessary to calculate
 c^n?fSS concentration for the fine and coarse particle fractions of a  dichotomous
 sample.  The following calculations are not required to determine attainment of the
           *ec?njar* standards; rather, they are  supplemental and may not be neces-
               - otomous sampler is designed to  fractionate a total PM10  sample" into
           ?nS12e hra:9eS  [Iine Par^cles  <1ess tha" 2.5 jun) and coarse  particles
    r*ia-  ? ^ ' ?f 9reater than 2.5 ^m)].  Note:  A correction is required for

 illustrated^               ^ P3rtl'CleS ComEted 0" the "'™ fil?er<
by Equation              Concentrat1on of the fine part id e fraction as calculated

                                   -  Mf(lQ3)
                                   '

-------
                                                            Section No.:  2.10.5
                                                            Date:   April 11,  1990
                                                            Page:   7
where
        [F] = mass concentration  of  fine  particles, /
-------

-------
                                                            Section No.:  2.10.6
                                                            Date:   April 11,  1990
                                                            Page:   1
 2.10.6  MAINTENANCE
      Preventive maintenance is  defined  as  a  program of  positive actions  aimed
 toward  preventing  failure of monitoring and  analytical  systems.  The overall objec-
 tive of a  routine  preventive maintenance program  is to  increase measurement  system
 reliability and to provide for  more  complete data acquisition.

      This  subsection  outlines general maintenance procedures  for a  specific  commer-
 cially  available dichotomous sampler.   For more complete  information on  a  particu-
 lar  sampler,  or on laboratory equipment maintenance,  refer  to the manufacturer's
 instruction manual  for  the individual instrument.

      Records  should be  maintained for the  maintenance schedule  of each dichotomous
 sampler.   Files  should  reflect  the history of maintenance,  including all replace-
 ment  parts,  suppliers,  cost- expenditures,  and an  inventory  of on-hand spare equip-
 ment  for each sampler.                                                 H     M  F

 6.1   Maintenance Procedures

 6.1.1   Recommended Supplies  for Maintenance  Procedures -

      An alcohol-based general -purpose cleaner, cotton swabs,  a  smaH soft-bristle
 brush,  paper towels, distilled water, and miscellaneous handtools are required
maintenance supplies for dichotomous samplers.  A compressed-air  source is recom-
mended-,  but not .required.

6.1.2  Sampl ing Module - "   -                                  •

     The sampling module of the  dichotomous sampler consists of the sampler inlet
and the  virtual  imp-action assembly.   Figure 6.1  shows, a disassembled sampling
sealed with "O^rin   ll1ustrates the virtual impactor assembly.  All parts are
                                                      NOT BE DISHANTLED-   CHECK
     To dismantle the sampler inlet:

          Mark  each  assembly point  of the  samp-ler inlet  with pen  or pencil  to pro-
          vide  "match marks" during reassembly.

     •  '   Disassemble the  unit  in accordance with manufacturer's  instructions
          taking  care to retain  all  "0"  rings  and miscellaneous parts.
          N£le:   If  the assembly screws,  appear frozen, the  application  of  penetrat
          ing oil  or commercial  lubricant  will make  removal  easier.

          Clean all  interior surfaces  with the general-purpose cleaner  or
        •  compressed-air source, paying  particular attention  to small openinqs  and
          crevices.   Cotton  swabs and/or a small  brush would  be most  helpful  in
          these areas.  Completely  dry all components.

-------
 Bug Screen
Lower Plate
with Rain
Deflector
      Collector
      Assembly
                                                          Section No.:   2.10.6
                                                          Date:   April  11,  1990
                                                          Page:   2
                                                                Top Plate with
                                                                Deflector Cone

                                                                Pan Head Screw
                                                                Spacer
                                                               Acceleration
                                                               Assembly
                                                            O-Rings
                        Figure 6.1.  Dichotomous sampler inlet.

-------
                                                          Section No.:   2.10.6
                                                          Date:   April  11,  1990
                                                          Page:   3
Receiver
Nozzle Assembly
     Coarse Tube

       Fine Tube
   Filter Cartridge



 Fine Filter Holder
          Bracket
                                                           O-Rlngs
                                                            Inlet Tube
                                                           Accelerator
                                                           Nozzle
                                                                 0-Rings
                                                         Coarse Filter
                                                         Holder
      Figure 6.2.  Dlchotomous sampler virtual impactlon assembly.

-------
                                                           Section No.:  2.10.6
                                                           Date:  April 11, 1990
                                                           Page:  4


     •    Reassemble the unit in accordance with the previously scribed match
          marks.  Take particular care to ensure that all "0" ring seals are
          properly sealed and that all screws are uniformly tightened.

     The "0" rings in the aerosol inlet should be removed periodically and
conditioned with vacuum grease.  This will inhibit breakdown and fraying of the "0"
ring caused by friction on the inlet tube.  The bug screen protecting the aerosol
sampler inlet should be cleaned periodically during the summer months.  The bug
screen is exposed for cleaning by pulling the sampler inlet.off the receiver tube
assembly.  An "0" ring in the sampler inlet acts as the seal.  Many samplers are
equipped with an inlet that also has a primary water trap on the exterior of the
unit.  If this trap is glass, care should be taken not to crack or break it, as the
sampler will not maintain adequate vacuum during operation.  The glass -trap may
either be replaced with a plastic jar or wrapped with insulating tape to minimize
the danger of accidental breakage.

Virtual Impactor Assembly - Internal particulate deposits accumulate primarily on
the outer and inner surfaces of the tip (closest to sampler inlet) of the inlet
tube.  Thus, the inlet tube should be inspected periodically for such particulate
deposits and cleaned as required.  An inlet tube cleaning schedule of every 3 to 4.
months is typical; the remaining inner surfaces should be cleaned every 6 to 12
months.  Use alcohol or water and a soft-bristle brush for cleaning.

   '  Examine sample module vacuum tubing periodically for crimps, cracks,  or breaks
and replace as necessary.  Examine connecting fittings for cross-threading, and
replace fittings if necessary.

6.1.3  Control Module

CAUTION:  UNPLUG THE POWER CORD FROM ITS RECEPTACLE BEFORE REMOVING OR OPENING THE
FRONT PANEL OF THE DICHOTOMOUS CONTROL MODULE.

Control Module Cleaning Procedures -

     1.   Remove or open the front panel and blow out loose dust and dirt if com-
          pressed air is available.   Wipe down all  surfaces with the general-pur-
          pose cleaner and towels.

     2.   Make note of any obvious problems in the unit and take action to correct
          them before completion of cleaning.  Refer to the manufacturer's instruc-
          tion manual.

     3.   Check rotameters for cleanliness.  If they are dirty and/or contain
          water,  they must be removed and cleaned.   (If water is found,  the inter-
          ior of the vacuum pump may be damaged.   It should be opened for inspec-
          tion and possible repair.)  To clean the rotameters,  take the following
          steps:

-------
                                                             Section  No.:   2.10.6
                                                             Date:  April  11,  1990
                                                             Page:  5


           •    Remove the tubing from the total  rotameter output port  and any
                other connecting tubing that may  prove too inflexible to  allow
                removal of the rotameters.

                Remove the screws securing the rotameter assembly to the  front
                panel.                                                     .

           •    Slip the assembly back from the front panel enough to gain
                access to the Allen screws in the top of each rotameter and
                remove the protective cover,

                While holding the glass rotameter with one hand, loosen the
                Allen screws just enough to allow removal  of each of the gradu-
                ated glass tubes.

                Clean the two rotameters with an alcohol-based cleaner and
                rinse thoroughly in  distilled water.  For  proper cleaning of
                the unit,  the float  and its retainers also' should be  removed
                The retainers are easily removed with the  aid of a wire hook
                fashioned from a paper clip.

                Allow the tubes  to' dry thoroughly and reassemble.              •

      4..   Remove  and clean  all  filter jars.   Check  each for  possible cracks
           and  replace if. necessary.   Should  a  filter jar  become cracked or '-
           loosened,  the  dichptomous  sampler  will  not maintain an  adequate
           vacuum  during  system  leak  tests.   Be certain that  each  filter jar  is
           tightened  and  sealed  properly.   Clean  or  replace any  dirty filter
           elements.   These elements  may become dirty in routine operation or
           •  ™«  samP'er  1S  inadvertently  energized without  sample filters
           installed.

      5.    Clean the  cooling  fan's blades an.d housing with compressed  air  or a
           sma.ll brush.  Check the housing  for  any dirt that  could cause the
           fan to  lock up.

      6.    Clean exterior surfaces of  the vacuum pump; be sure that all cooling

           rhlrl 3n  °Pen:- Tat6 Care  that f1uids do  not run  inside the P"«np.
           Check all mounting brackets to ensure that  they are tight and in
           good condition.                     .                 •

Vacuum Pump - It  is  recommended that  the diaphragm  and the flapper valves of
                 ? rep1aC6d routine1y 
-------
                                                           Section No.:  2.10.6
                                                           Date:  April 11, 1990
                                                           Page:  6


the procedure, making sure that the screw clearance cavity in the.plate is
lined up under the intake valve screw heads.  All head screws must be tight-
ened evenly.  Diaphragms are often available through local suppliers or they
may be purchased through the manufacturer.

     When all cleaning and routine maintenance operations have been 'completed,
close the control module, reassemble and connect the sample module, and recal-
ibrate the instrument (if necessary).  Refer to Subsection 2 for calibration
procedures.

6.2  Refurbishment of Dichotompus Samplers

     Dichotomous samplers that have been operated in the field for extended
periods may require major repairs or complete refurbishment.  In these cases,
the manufacturer's instrument manual must be referred to before work is under-
taken.  A dichotomous sampler that has been subject to major repairs or refur-
bishment must be leak-checked and calibrated prior to sample collection.

-------
                                                            Section No.:   2.10.7
                                                            Date:   April  11,  1990
                                                            Page:   1


 2.10.7  AUDITING PROCEDURES

      The operating agency must perform QA audits and process  evaluations  to
 determine the accuracy of the PM10 monitoring system and,  hence,  the  data  it
 produces.  The primary goal of an auditing program is to  identify system  errors
 that may result in suspect or invalid data.   The efficiency of  the monitoring
 system (i.e.  labor input vs. valid data output)  is  contingent  upon effective QA
 activmes:  This true assessment of the accuracy and efficiency  of the PM10
 guide'hnes: **     "" °"ly ^ achieved b* conducting an audit  under  the following


      •     Without special  preparation  or adjustment  of the system  to be audited.

           By an  individual  with  a thorough  knowledge  of the instrument or process
           being  evaluated,  but not  by  the  routine operator.               process
           rnmnt-              . NIST-traceable transfer standards that are
           completely  independent of those used for routine calibration and QC flow


           With complete documentation of audit information for submission to- the
           operating agency.  The audit -information includes,  but is  not  limited to
                                nd   11'* tr™**r standards  instrument  rnoSe  and  '
                                            t™"b'"ty,. calibration  inflation,
estimates™?^ K??S£?ad"CHb?i in thi$ subsection produce  two  quantitative
estimates of a PM10 sampler's performance:   the audit flow rate oercentaoe
oercenlace6 Tiffl^W f1°W rat* Percentage difference.   ThI KdU  flow  rate
percentage, difference determines the accuracy of 'the  sampler's  indicated flow  ratP
by comparing it with a flow rate from the audit transfer  standard   The des?on ??
SieiS?rJe3ta?e difference determines how closely  t^sample^s  ?iow rate mltche
the inlet design flow rate under normal  operating  conditions.             matches

     An independent observer should be present for the audit  oreferablv thP
tSfJntecrufo0? ll^u^l^ ««uiP-?nt-   This  Practice not^nly'coni? utes to
 he            -6     *
                                                           Poscsso
discrepancies between  audit-standard  values and the lamPHngPequ!pmen"values[

                          di:fferences in flow rate (between audit flow and sampler
                          ling equipment malfunction or operator technique  an
                                                                     '
          Flow-rate performance audit.

          System audit of data processing.

-------
                                                           Section No.:  2.10.7
                                                           Date:  April  11, 1990
                                                           Page:  2
          Analytical process system evaluation.
     These audits and evaluations are summarized in a table at the end of this
subsection (Table 7.1).  Refer to Section 2.0.11 of this volume for detailed proce-
dures for systems audits.

     Proper implementation of an auditing program serves a twofold purpose:  to
ensure the integrity'of the data and to assess the accuracy of the data.  Addition-
al information on assessing the accuracy of the data is given in Section 2.0.8 of
this volume of the Handbook.

7.1  Flow-Rate Performance Audit

     The following subsection presents audit procedures specific to commercially
available dichotomous samplers which operate at an actual total flow of 16.7 L/min
and a coarse flow of 1.67 L/min.  Audit techniques may vary between different
models of samplers due to differences in required flow rates and the sampler's
sampling configuration.

     The dichotomous sampler flow rate audit method involves using two transfer
standards.  One is calibrated in the flow range of the total and fine flow rates
and the second is calibrated within the range of the coarse flow rate.  This
enables the auditor to measure the critical flow rates directly without compounding
transfer ..standard error through subtraction.  Obviously, the optimum audit method
would incorporate one transfer standard calibrated over the entire range of the
sampler's accepted flow limits (1.5 L/min to 18.4 L/min).  Accuracy over this flow
range is difficult to achieve within acceptable limits.  Consequently, it is
recommended to conduct audits using transfer standards withrn -specific ranges to
measure the sampler's indicated flow rates. .

     Since the accurate measurement of PM10 mass concentration is dependent upon
flow rates under actual conditions, the auditor must also audit in terms of actual
conditions.  If the audit transfer standard's calibration data have been corrected
to EPA reference conditions (298 K, 760 mm Hg or 101 kPa),  a conversion must be
calculated to adjust the standard L/min flow rate (Qstd) to an actual L/min flow
rate (Qa).

7.1.1  Audit Apparatus -

     Any type of flow-rate transfer device acceptable for use in calibration of
dichotomous samplers may be used as the audit flow-rate reference standard; how-
ever, the audit standard must be a different device from the one used to calibrate
the sampler.   The audit standard must be calibrated against a primary standard
traceable to the MIST.   Refer to Subsection 2,  Tables 2.1 and 2.2,  which reference
flow-rate transfer standard calibration procedures.  Assemble the audit apparatus
as indicated in Figures 7.1 through 7.3.  In addition to that which is presented in
the tables, a few miscellaneous supplies are required.   These include a 9.53-mm
(3/8-in.)  Swagelok cap,  6.35-mm (1/4-in.)  Swagelok cap,  hand tools, and an adapter
to connect the transfer standard outlet to the sampler inlet.

-------
                                                  Section No.:   2.10.7
                                                  Date:  April  11,  1990
                                                  Page:  3
       Audit
     Adaptive
      Devic*
 Chosen
 Transfer
Standard
                             Coarse Row
                                       0  0
                               Coarse
                                Flow
                             Rotameter
  Total
  Flow
Rotameter
 Figure 7.1. Audit assembly and dlchotomous sampler
          set up to audit total flow (TQa).

-------
                                                 Section No.:
                                                 Date:   April
                                                 Page:   4
              2.10.7
             11,  1990
        Audit
       Adaptive •
        Device
 Chosen
 Transfer
Standard
                              Coarse Flow
                                       Disconnect 6.35 mm (1/4 In) O.D.
                                       tubing and install Swagelok cap

                                         Paniculate filter installed on
                                         6.35 mm (1/4 In) O.D. tubing
                                 =    00
                                Coarse
                                 Flow
                               Rotameter
  Total
  Flow
Rotarneter
 Figure 7.2. Audit assembly and dichotomous sampler
            set up to audit fine flow (FQa).

-------
                                                                    Section No.:   2.10.7
                                                                    Date:   April  11,  1990
                                                                    Page:   5
                           Audit
                          Adaptive •
                           Device
                     Chosen
                     Transfer
                     Standard
                              Fine Row
Coarse Flow
 Disconnect 9.53 mm (3/8 in) O.D,
 tubing and install Swagelok cap
Paniculate filter installed on
9.53 mm (3/8) O.D. tubing
          0   0
                            nrm
                                                    Coarse
                                                     Row
                                                   Rotameter
                 q    Total
                      Row
                   Rotameter
                     Figure 7.3. Audit assembly and dichotomous sampler
                              set up to audit coarse flow (CQa).

-------
                                                           Section No.:  2.10.7
                                                           Date:  April 11, 1990
                                                           Page:  6


     An audit data sheet similar to Figure 7.4 must be used to document audit
information.  This information includes, but is not limited to, sampler and audit
transfer standard type, model and serial numbers, transfer standard traceability
and calibration information, ambient temperature and pressure conditions, and
collected audit data.

7.1.2  Total Flow-Rate Audit Procedures -

     1.   Instruct the operator to install new filters .in both the fine and coarse
          filter holders and energize the sampler.  Filters used for flow rate
          audits should not be used for sampling.

     2.   Instruct the operator to adjust the rotameter flow-control valves to set
          the total and coarse rotameters to their operational set points for rou-
          tine sampling.  These set points should correspond to the calculated set
          points (TSP, CSP) determined, by the sampler's calibration relationship.

     3.   Allow- the sampler to warm up for a minimum of 5 min while maintaining the
          proper total and coarse rotameter set points.

     4.   Complete the top half of the data sheet with the required information,
          including ambient temperature (Ta) and ambient barometric pressure (Pa).
          Record both the TSP and CSP values and the corresponding flow rates.

     5.   Remove the sampler inlet and replace with the transfer standard adaptive
          device (see Figure 2.1).

     6.   Connect the adapter to the transfer standard outlet with flexible tubing,
          being careful not to crimp the tubing.  If the transfer standard is elec-
          tronic,  it must equilibrate to operating conditions.  A warmup time of at
          least 5 min is recommended.

     7.   Recheck rotameter settings;  if different from designated .set points,
          record new value and the-corresponding flow rate as determined by the
          sampler's calibration relationship.

     8.   Record on the audit data sheet the transfer standard (TS) readings
          (volts,  AH20, timings,  etc.).

7.1.3  Fine Flow Rate Audit Procedures -

     1.   Turn the sampler off and disconnect the coarse-flow 6.53-mm (1/4-in.)
          line.  Cap the coarse-flow outlet port located beneath the dichotomous
          sampler filter holder with a 6.53-mm (1/4-in.)  Swagelok cap.  This opens
          the coarse line to the vacuum pump.  To prevent particle entrapment
          within the system,  it is recommended that a particle-free filter be
          attached to the line.

-------
                                                             Section  No.:  2.10.7
                                                             Date:  April 11, 1990
                                                             Page:  7
Dlcnotomous Sampler Audit Data Sheet
Station Location L&^f /L/£L natft ^ /// /£*? q * no A n M, ,mh nr *1 ^ ^ ^1-1 <&
Station Address /O / /O. f/5

3^
 Transfer Standards Calibration Relationships:
 Total and Rne:  m . 23.
-------
                                                           Section No.:  2; 10-.7
                                                           Date:  April 11, 1990
                                                           Page:  8


     2.   Turn the sampler on and check the rotameter set points.  If variation has
          occurred since the total flow rate audit, record the to'tal and coarse
          rotameter units and their corresponding flow rate values determined from
          the sampler's calibration.  A small flow imbalance occurs when the coarse
          line is disconnected; this may cause rotameter fluctuations.

     3.   Record on the audit data sheet the transfer standard (TS) readings
          (volts, AH20, timings, etc.).

7.1.4  Coarse Flow Rate Audit Procedures -

     1.   Turn the sampler off and exchange the total and fine flow rate transfer
          standard for the coarse flow transfer standard.  If necessary, allow this
          transfer standard to equilibrate to ambient conditions (at least 5 min).

     2.   Reconnect the coarse flow line and disconnect the fine flow 9.53-mm (3/8-
          in.) line.  Cap the fine flow outlet port located beneath the dichotomous
          sampler filter holders with a 9.53-mm (3/8-in.) Swage'lok cap.  This opens
          the fine line to the vacuum pump.  To prevent particle entrapment within
          the system, it is recommended that a particle-free filter be attached to
          the line.

     3.   Turn the sampler on and check rotameter set points.  If variation has
         -occurred since'the total flow rate audit, record the total and coarse
          rotameter units and their corresponding flow rate.values determined from
          the sampler's calibration.  A small flow imbalance occurs when the fine
          line is disconnected; this may cause rotameter fluctuations.

     4.   Record on the audit data sheet the transfer standard (TS) readings
          (volts, AH20, timings, etc.).

7.1.5  Audit Data Calculations -

     1.   Calculate and record the audit total, fine, and coarse flow rates by.
          using the calibration curve accompanying the transfer standard.  Record
          these values to the nearest 0.01 L/min (e.g., 1.67 L/min) on the audit
          data sheet.

          Note;  It may be necessary to correct audit flow rates to actual condi-
          tions1.  If a1 soap film flowmeter has been used to determine the coarse
          flow rate, no water .vapor corrections are necessary for this audit flow.
                            Qa = Qstd(Ta/Pa) (Pstd/Tstd)
                                                                 (Eq: 1)
where
     Pstd,
  Qa
Qstd

  Ta
  Pa
Tstd
flow rate at actual conditions, L/min
flow rate corrected to standard  temperature and pressure (25
298 K; 760 mm Hg or 101 kPa),  L/min
ambient temperature, K
ambient barometric pressure, mm Hg or kPa
standard barometric pressure and temperature, respectively.
                                                                                 'C,

-------
                                                              Section  No.:  2.10.7
                                                              Date:  April  11, 1990
                                                              Page:  9
       2.   Instruct the operator to calculate (using the sampler's calibration rela-
            tionship) the corresponding sampler flow rates and record.

       3.   Determine the percentage difference between the sampler-indicated flow
            rates and the audit-measured flow rates as:


            % Difference = ^^(Ldi?) ^-1 (100)                     (Eq.  18)


                                                b*tween the sa*Pler <«'*» flow rates
            % Difference  =  Qa (audit)  - Design flow rate
                                  Design  flow rate                           (EQ-  19)
     T?T?:| /m?n I"?** C^m?c ^'fhotomous samplers, the design flow rates are
     conditions.       '        /min ( ine)> and K67 L/min (coarse) at actual


5'   ^°rJhSerCenf diff?r;nce:  If the deference is less than or equal  to
          rp  ,Slrr,oali •KatJr.  the audit should be

 7-1-5  Performance Audit  Frequency -










7.2   Systems Audit






procedures to evaluate data processing anS labo^tory operations "  SyStem$ 3Udlt

-------
                                                           Section No.:  2.10.7
                                                           Date:  April 11, 1990
                                                           Page:  10


     Subsections 2.0.11 and 2.0.12 of this volume provide detailed procedures and
forms for systems audits and performance audits, respectively.

7.2.1  Systems Audit of Data Processing -

     It is recommended that data processing be audited soon after the original
calculations have been performed.  This allows corrections to be made immediately
and also allows for possible retrieval of additional explanatory data from field
personnel when necessary.  A minimum frequency of seven samples per 100 (minimum of
four per lot) is recommended.  The procedure is as follows:

     1.   Use the operational flow rates as reported on the-sample data sheets.

     2.   Beginning with the raw data on the dichotomous sample data sheet and the
          filter net- and tare weights, independently compute the concentration
          (^g/std. m3) and compare it with the corresponding concentration origin-
          ally reported.  If the mass concentration computed by the audit check
          (/jg/std. m^) does not agree with the original value within round-off
          error, recheck all samples in the lot.

     3.   Record the audit values on a data sheet, and report them, along with the
          original values, to the supervisor for review.  The audit value is always
         ..given as the correct value based on the assumption that a discrepancy
          between the two values is always double-checked by the auditor.

7.2.2  Analytical Process System Evaluation -                                   .

     A performance audit of the microbalances used to weigh dichotomous filters
would require the use of ASTH Class 1 standard weights.  Since microbalances are
extremely delicate instruments and should not be operated by inexperienced person-
nel, it is recommended that the performance evaluation of the filter weighing proc-
ess be done in the following manner:

     1.   Review the maintenance and calibration log for each balance.  Routine
          balance maintenance and calibrations must be performed by the manufactur-
          er's service representative at manufacturer-specified scheduled inter-
          vals.  In no case should the interval between calibrations exceed 1 year.

     2.   Review QC data records for the filter-weighing process.  Ensure that the
          following QC activities have been performed and documented:

          •     Zero and calibration checks after every five filter weighings.

          •     Standard filter weighing every day of the balance operation.

          •     Duplicate filter weighing for every five to seven filters.

          If QC checks were out of limits, note what action was taken.

-------
                                                  Section No.:  2.10.7
                                                  Date:   April 11,  1990
                                                  Page:   11
      Select  randomly  and  have  the  balance  operator reweigh  four equili-
      brated  filters out of  every group  of  50  or  less.   For  groups  of  50
      to  100,  reweigh  7 from each group.  It is of  primary Importance  to
      be  sure  that the sample is representative of  the  various  conditions
      that may influence data quality.

      Record the original  values and the audit weights  on the audit form.
  ,    Calculate the weight difference for each filter as follows:

Difference = Original  weight (mg)  - Audit weight (mg)

For unexposed filters, the difference should be less than ±20 itg (0.020

h?b,:t< JrreTSed/fl  ter-' *?? P°tentl'a1 loss  °f volatile particles pro-
Sltl to .H Pf KCe/rejeCt1°n  limits  to be established.  Forward the aSdit
data to the laboratory supervisor  for review.

-------
                                                                     Section  No.:   2.10.7
                                                                     Date:   April  11,  1990
                                                                     Page:   12
j  •

4>  *1
U  C
lit
       T)
        P
       T>
        c
       a.
        U
        O

       0.
          •   U I
          U JC C —
        I  C O * — •
        E • • k •
        3 k X • >
        • • U *- C

        k t 4 i *j

        «-52"° x"
        k   3*- B
        O %- O HI iH
       *- —-O   +1.
                                  Jf

                                  e
              P — +1— O
             k O.
             x- e
                 • •
             • C X • 4>
             cc — • u • •
                   t.
                   o
               §3
             *)


             X- *- • X


             • TJ f **l
             O • I
             C 4> *> C


             k S-D X
             • — 3 4>
             &-!„
             ^i^s
             lll
             •
             0. *>
            o •
            V) X
       58
       k C
       • o
            X C
             o —
             • k
             • o
             o c -e
             o
            u.
 _ o
C£ —
                              -o •
                              • o
                              • a
                              O X
                            — o. •
                             • X
                             C • k
                            — CO
                             a 3 %-
                                 X   O
                                  as c
4> X •

if=
< *••-
                                               u
                                             — c
•f*

li
            •o —
            c •
            • 3
              C
            •o c
                                             e •
                                        • L.
                                        m jj
                                   s
                                 8
                                      O
                                      t
                                      U
                                        3 U
                                        o- c

                                        C C

                                    *   -55

                                    £   5JE
                                    X     f
                                    ai   at 6
                                                     4>


                                                     "3
JC

J

U

£
          i


        $5

        ?§'t
        QUO
        •   I

        C k C
        003
        — a o
        C • X'


        c 'ou"»

        U k C
          o o •
        *>   — k
        — X 4> O
        •o *> • k
        3 — k k
        <••»>•
                                                     U O • k
                                                     u • r»

                                                     •*> *> k *
                                                     • c o •»
                             T)  •


                             JI
                              E  O

                             -o  a.

                              >  9

                              2.5
                                                               • k

                                                              2^.
                                                              — •«
                                                               c c
          X C
          *> o
                                                         C 4>
                                                         — o
                                                         J
                                                                x o
                                                                 3 O
                                                         k   -o
                                                         o -e c
                                                         • CM
            C •
         •oo —
          O — X
         X 4> *»
         *> u

         ^:^
                             c — u
                             I 8 kJ
                                             X O —
                                             an c
                             C — > Jt
                               k k U
                             • f • •
                               6. • x
                                 JO U
                               •o o
                                   .O 3
                                 *. J S -O
                            I
                            •
                            3
                            >
                       U

                      *j
                            • o

                            < '••>
                                  k •

                                 a.'
                                  c

                                 X
                                  a
s
        • c -o e
        k o k 5  %-
          u o E   o
                                                c *•  • 5   — • k
                                                • o -D S   a x a.
            • — —   • C E
            «>—  -Q05k     	
            -^•«   C— k*   4>C&3
            OKU  t-l *> S- O.   < — • -O
                                                              .
                                                             o
                                                             k •.
                                                           a a u
                                             U
                                             c
        u
        o

        a

-------
                                                            Section No.:  2.10.8
                                                            Date:  April 11, 1990
                                                            Page:  1


 2.10.8  ASSESSMENT OF MONITORING DATA FOR PRECISION AND ACCURACY

 8.1  Precision

      One or more monitoring sites within the reporting organization are selected
 tor duplicate collocated sampling as follows:  for a network of 1 to 5 sites   1
 site is selected; for a network of 6 to 20 sites,  2 sites are selected; and for  a
 network of more than 20 sites, 3 sites are selected.  Where possible,  additional-
 collocated sampling is encouraged.   Annual mean particulate matter concentrations
 of the selected sites should be among the highest  25% of the annual  mean PM10
 concentrations for all the sites in the network.   If such sites are impractical
 however,  alternate sites approved by the Regional  Administrator may be selected!

 aii   K0l12CaJed PM1° samPlers  bein9 "sed for assessment  of precision should gener-
 ally be of the same type.   That is,  they should have similar flow rates (e  q
 c^iJ^r"1'  orj°?!' similar inlet types (e.g.,  impaction or  cyclonic), and'
 —"i- flow controller types  (e.g.,  MFC or rotameters).   Where a PM10 network
 anar+Tl!e  tw°  Allocated  samplers must be within 4 m of each other, but at least 2 m
 bS  tL  LJTS  dK  f;r  f  ?W  intrference'  Calibration, sampling, and analysis must
 on. nf    K   01- b°Jh C?  10Cated samP1ers and all other samplers in the network.
 wh?c£ «JSuEai-i?fhCOll°Saied 5amplerS is desi9nated as the primary sampler from
 SJ2 If?^8/'1]   !  US6d  ?° rep°rt ai> qualit>' for the site; the other is desig-
 nated as  the  duplicate sampler.  Each duplicate sampler must be operated concur-
 ^^1^5 ^ associated  routine sampler at least once a week.  The operation
 the  tlr  !n2    ^tJ61?^6*1 S0,that the ™m?]^ **** *™ distributed evenly oJer
 the  year  and  over  the  7  days .of the week.  The every-6th-day schedule used by many
 Si  o? J3 a9en^es is recommended.  The measurements from both samplers at each *
          •    rP-lin9 l^* *£ reP°rted.  An example precision data reporting form is
           ?  hTre 8;J- Jhe P?^centa9e differences in measured concentrations
8.2  Accuracy
asseJJrf h!;C^MH5t<0f l^ dl'c5otomous sampler method in the measurement of PM10 is
assessed by auditing .the performance of the sampler (at its. specified flow rate)  as

flow'ate are'renortlS" ' ?1  ^ ^ 3Udit ^-"te and the corresponding samp er
F cure 8 2   Thfnlrrit,   H^6 accuracy data reporting form is presented in
      c u acyT            6i  \f ^ll

-------
                                                                 Section No.:

                                                                 Date:   April

                                                                 Page:   2
                                                                                 2.10.8

                                                                                11,  1990
2
g

GO

o
UJ
cc
a.
s
as
O

o
z

-
z
UJ

CO
CO
UJ

CO
<

O

a

3























I
|
s
•
!
f
S
*
£
T


i
i
1
•;
?
i


Tl.t,
tttl



•
<
£
, O
*•

c
c |
§

g
> 1

NOUVZINVDy
ONI1U041M
0
M
S










0
o
u
VI
a
O

1


,.





si
s:
tc
MM
a




q-
j-

•
«

—
-j




















*








•

•


N
•

•
r








Ml




























i
§
s
a
S
1
M.
0
5
X









o
QUJ
10
•- o
UJ (J

X
5











^
o

Ifc
c







	 aiiim
S
^
a


































5
>
a
X
O








§
i



-
«







r.

05'
0

S 


w- IU  **•
< ri




a
O ai <•>
X Q €N
>- O —

D I-
P



<02
O u T
OT K ""
































































































































































































































































































































































.














































a--








































0.








































































































































































































































































































































































































































                                                                                          05
                                                                                          CO
                                                                                          u
                                                                                          CO
                                                                                          05
                                                                                          a
                                                                                          cr
                                                                                          a
                                                                                          •5
                                                                                          •o

                                                                                          _o
                                                                                          w)
                                                                                         eo

                                                                                          £


                                                                                         i?
                                                                                         u.

-------
Section  No.:   2.10.8
Date:  April  11, 1990
Page:  3
oiiruo
oisiAiM
Y1T9IMO
•«••
—
—

LJ
u








— 1







—

— 1
^
iU





—
—

— J
s




••

—

— J
>
w








—1
?






—

__
>
w






^^v

M
a






—

__
>




—

—

-
             O Ml
             »0
             OE

             §!
             «: <

             II
             uo
             §s

             "
                        .0


                        'c
                        « •


                        o)
                        (fl
                        
-------

-------
                                                            Section No.:   2.10.9
                                                            Date:  April  11,  1990
                                                            Page:  1


2.10.9  RECOMMENDED STANDARDS FOR ESTABLISHING TRACEABILITY

     Two factors are essential for attainment of data of the desired quality:
(1) the measurement process must be under statistical control at the time of the
measurement, and (2  the combination of systematic errors and random variation
^measurement errors) must yield a suitably small  uncertainty.  Evidence of qood
quality data requires the performance of QC checks,  independent audits of the meas-
urement process, careful documentation of data,  and  the use of equipment and
instrumentation that can be traced to an appropriate primary standlrd

     The following  standards are recommended for  establishing tr-aceability:

          ASTtt Class 1  weights are recommended for the laboratory  microbalance
          checks3     '       Subsection 4.5  for details on balance  calibration


     2.    A  positive-displacement  primary"standard or laminar flow element is
          recomi;,endedflfoLcalibrating_the_f;ow-rate  transfer standard  that is used
     ?hS tit  <~   flacement primary standard is recommended for calibrating
     the transfer standard used to audit the dichotomous flow-rate calibra-
 -  ' audits.    Subsectlon 7A for details on the flow-rate performance


4.   The elapsed-time meter should be checked semiannuaHy against an accurate
     timepiece,  and it must be accurate within 15 min/day.             accurate


5'   barometeVf^toV^^                              (i'6"  then?onieterf•
     and against standards of known accuracy and traceable  toVsT.
                                                                               3

-------

-------

-------
 'i'  .',, :i ? DI '' 4'''ji % ir i: i' r; "i	f 11< i1' w'i nil	lit -i: 'ii'iiiiihiiH^	:nu»lii niii'ii iti i' iiiWM^

-:•• :>-vIi 'i«iI'"i'':1'i-i;![ "'T'ii f5iiliin^                                      Il'l''ivfiK!"tiii:,"••SiyiiSv.i!iiiii!iiiiNilil
             l • i  , I
                                 ji' M'jii  iiiiiim r	in miiivni!,';1 iiiii|iiiiiinniiiiii' ill1' ii	,i f ii'i' iH'i	le is 1.1,	/, tn i' iiiiii: ii: ii i,:: v: j "i "i I1,: ii«' ii" J ill	i' «! J"., •' i' iiiinti "Viiiiwinir ii'iiifiiiiiir; •! 11 iin i,:> iiiiiir
                                                                                                      Ililllllilll 11 IP    II  II 11  III III
                                                                                                      III 111 III 111  1(11 II   I   II
                                                                                                         II   111   II
                          	I  ,1" "I „,' "I' i'i|| »,i|l' Jif!', llllilljll ,'l' ,1,1,,:" "I i liini Ml1 ft1	Illiililliiiil il' r "I I1'!"!1,1 Il'l it i il II llll111 iil'llll1	1 rill'li" Nil' il	''. >nii! ''Ii, 11 II  li'l  . I	|: I," ii	'"V', I,	Ill' i|ll' II iMIII'I'ilil'llii1,


                         ;i!ifi!!i^
                                                                           ffi!;*lKjS!i;';S(!r;!;| •'^SJ^!««iraS1B	I'll
                                                                           "Ll""	'	'	'	''ill
                               '"I.i11"1,1! iJI, 'IIiiSM1''!,:'!!",;!!!};! B'MBBik'SHS'.l 'i/'lll'iiE.!!'!..^!!!"!''!!!!1';' I!",;li1l .J'l 'I'l...'.]., '• it;1	!'; i !b;	MSWlf •IWlS't.iM**!
;i ' 'ill tip , ''iiiitiilii'
i jiv :*• jtfjsi, ...... ,:!::; if
                                                                                                                     ::^

-------
                    Addendum to Section 2.11
 Reference Method for the Determination of Paniculate Matter as PM10
       in the Atmosphere (High-Volume PM10 Sampler Method)
This section is up-to-date except that brushless motors are now available
for  high-volume  PM10 samplers.   According to their vendors,  these
motors have a maintenance-free operational life of 20,000 hours.  Also,
the  blank data  forms that are mentioned in. this section  have been
removed.

-------

-------
SECTION 2.11
                         Part1cuute Hatter
                                                                pmo
          Sub-
         section

          0
          0
          2
          3
          4
          5

          6
          7
          8
         10
         11
         12
                                       Outline
          Title
 Introduction
 Method Highlights

 Procurement  of Equipment  and
   Supplies
 Calibration  Procedures
'Field  Operations
 Filter Preparation  and  Analysis
 Calculations,  Validations,  and
   Reporting  of  PM10 Data
 Maintenance
 Auditing Procedures
 Assessment of Monitoring  Data
   for  Precision and Accuracy
 Recommended  Standards for
   Establishing Traceability
 Reference Method
 References
Data Forms
Full  Section
   Number

   2.11.0
   2.11.0
   2.11.1
   2.11.2
   2.11.3
   2.11.4

   2.11.5
   2.11.6
   2.11.7

   2.11.8

   2.11.9
  2.11.10
  2.11.11
  2.11.12
No. of
pages

   6
   2
   5
  30
  31
  10

  10
   5
  19
  1
  5
  1
 16
                                                        Printed on Recycled Paper

-------

-------
                                                               Section  No.: 2.11.0
                                                               Date:  January 1990
                                                               Page:  1
 2.11.0.  INTRODUCTION
 ao™H PM10.1S *he designation for particulate matter in the atmosphere that has an
 aerodynamic diameter of 10 micrometers  (H or less.  A high-volume  (HV) PM10 sam-
 ^"t-^5-3!^0^/^""16 uf ambient  air at a constant flow rate through a size-
 selective  inlet and through one or more filters.   Particles in  the PM10 size ranae
 ?£h  SJlS^u"1 °" th-  ri*ris) dun'ng the 5Pecif^d 24-hour samp ?ng peri o?9
 Each  samp  e filter is weighed before  and after sampling to determine the net weioht
 (mass) gain of the collected PM10 sample.  The referent method™? PM10 slmpTinq

           nFqUlatinnS(4° CFR 5°  Aendi              11"9
                                                                         .ii,
 computed as the total  mass  of collected particles  in  the PM10 size ?angfdiv  ded bv
                                                                s isrs
tiveness)  of the  sampler inlet over the PM10 size range.  Of particularP moortlnce
is the^particle size at which the sampler  effectiveness is 50 percent  (i e   ?he
SS inS «S1Ze SUt|?01nt)-  Methods for PM1° that »eet all requirements  n'both
50 and 53  are designated as PM10 reference methods for use in State  and Lorll
Monitoring System  (SLAMS) and Prevention of Significant olterioraJion  (PSD) m

f^urer I^lhf^f ol T^                             ™

      are discussed in this section; dichotomous sampled  are discussed in Section


-------
                                                                 Section No.:  2.11.0
                                                                 Date:  January 1990
                                                                 Page:  2
 Impaction  Inlet
      Figure 0.1  is  a schematic drawing showing the basic elements  of an  impaction
 HV PM10  inlet.   The symmetrical  design of the inlet insures  wind-direction  insensi-
 tivity.   Ambient air that is  drawn into the inlet is evacuated  from the  buffer
 chamber  through  nine acceleration nozzles into the first impaction chamber,  where
"Initial  particle separation occurs.   The air is then accelerated through an  addi-
.tional 16 jets  into a second  impaction chamber.  The acceleration  jets have  criti-
 cal  diameters calculated  by the manufacturer to provide  the  necessary changes in
 velocity to effect  correct particle  size fractionation within the  impaction  cham-
 bers.  The air  flow finally exits the inlet through nine vent tubes onto a sample
 fiHer.   Because air velocities  are  critical  to maintain the correct particle size
 cutpoint within  the inlet,  maintaining the correct design flow  rate through  the
 inlet  is important.  This design flow rate is specified  by the  manufacturer  in the
 instruction manual.  For  example,  the design  flow rate for one  popular impaction
 inlet  is 1.13 m3/min.                                                              <

 Cyclonic Inlet

      Figure 0.2  is  a schematic drawing of a cyclonic HV  PM10 inlet.   The omnidirec-
 tional cyclone used for fractionation in this inlet allows particles to  enter from
 a>TT  angles of approach.   Ah angular  velocity  component is imparted to the sample
 aitr  stream and the  particles  contained in it  by a series of  evenly spaced vanes.
 Lsirger particle  removal occurs in an inner collection tube.  This  tube incorporates
 a  "'perfect absorber"--an  oil-coated  surface to eliminate particle  bounce and reen-
 trainment.   The  sample flow (with the unremoved smaller  particles)  then  enters an
 frrtermediate tube,  where  the  trajectory is altered to an upward direction.   An
 additional  turn  is  then made  to  alter the flow to a downward trajectory  to allow
 the  remaining particles (i.e.,  PM10  fraction)  ultimately to  deposit  on a filter for
 subsequent analysis.  As  with the impaction inlet,  control of air  velocities in the
 cyclonic inlet is critical  to maintain the correct particle  size cutpoint.   Main-
 taining  the correct design  volumetric flow rate through  the  inlet  is  important.
 "WHs design flow rate is  specified by the manufacturer in the instruction manual.
 For  example, as  in  the case of the impaction  inlet,  a popular cyclonic inlet also
 Iras,  a design flow rate of 1.13 m^/min.

 Wass-Flow-Control'(MFC) System

     The flow rate  in a MFC system is actively  sensed and controlled  at  some prede-
 termined set point.  Air  is .pulled through  the  filter into the  intake of a blower
 and?  subsequently exits the  sampler through  an  exit orifice,  which  facilitates meas-
 urement  of  the flow with  a  manometer or pressure  recorder.   The flow rate is con-
 troTled  by  an electronic  mass-flow controller,  which uses  a  flow sensor  installed
 below the  filter holder to  monitor the  mass flow  rate and  to control  the speed of
 the-motor  accordingly.  The controlled  flow rate  can be  changed by  an adjustment
 knob on  the flow controller.

 Vtolumetric-Flow-Control (VFC)   System

     A VFC  system maintains a  constant  volumetric  flow rate  (given a  fixed tempera-
 ture) through the inlet,   rather  than  a  constant mass flow  rate as  in  the  MFC

-------
                                                       Section No.:  2.11.0
                                                       Date:  January 1990
                                                       Page:  3

                                                     Buffer Chamber




                                                     Air Flow

                                                     Acceleration Nozzle



                                                    Impaction Chamber


                                                     Acceleration Nozzle



                                                     Impaction Chamber

                                                     Vent Tubes



                                                    Filter Cassette

                                                    Fitter




                                                    Rlter Support Screen
                                                    Motor Inlet
Figure 0.1. Schematic diagram of an impaction inlet.

-------
                                                                           Section  No.:  2.11.0
                                                                           Date:  January 1990
                                                                           Page:  4
                                                Maintenance Access Port
                                         Perfect
                                        Absorber
                                        No-Bounce
                                         Surface
Housing-Deflector
    Spacing
                                                                         Vanes
                                                                          Wane
                                                                        Assembly
                                                                          Base

                                                                         Insect
                                                                        Screen
                                                                         Protective
                                                                         Housing
       Aerodynamic
           Inlet
         Pathway
Aerodynamic Flow
    Deflector

Outer Tube
                   Figure 0.2. Schematic diagram of a cyclonic inlet.

-------
                                                                Section No.: 2.11.0
                                                                Date: January 1990
                                                                Page: 5
system.  In a popular commercial VFC system, a choked-flow venturi is operated such
that the air attains sonic velocity in the throat of the device.  In this "choked"
mode, the flow rate is unaffected by downstream conditions such as motor speed or
exit pressure, and is a predictable function of upstream conditions,  such as the
stagnation pressure ratio and temperature.  Thus, the volumetric flow is controlled
without any moving parts or electronic components.  In this type of flow control
system, ,no means is provided for adjusting the controlled flow rate.

-------
                                                                 Section  No.: 2.11.0
                                                                 Date:  January  1990
                                                                 Page:  6
METHOD HIGHLIGHTS
     The procedures set forth in this document are designed to serve as guidelines
for the development of quality assurance  (QA) programs associated with the opera-
tion of an HV PM10 sampler.  Because recordkeeping is a critical part of QA activi-
ties, several data forms are included to  aid in the .documentation of data.  The
blank data forms (see Subsection 12) may  be used as they are, or they may serve as
guidelines for preparing forms more specific to the needs of the individual moni-
toring agency.  Partially filled-in forms are included at appropriate places in the
discussion of the procedures to illustrate their uses.

     Tables at the end of some subsections summarize the material covered in those
subsections.   The material covered in the various subsections of this section is
summarized here:

     1.    Subsection 1,  Procurement of Equipment.and Supplies,  includes recommended
          procurement procedures,  equipment selection criteria,  and minimum accu-
          racy requirements.  It also provides  an example of a  permanent  pro-
          curement  record.

     2.    Subsection 2,  Calibration Procedures,  provides  detailed calibration
          procedures for the HV  PM10 sampler.   A table at the end of this  subsec-
          tion summarizes  acceptance limits and  gives references  for the  calibra-
          tion procedures  of associated  monitoring  equipment used in a  PM10 sam-
          pling  program.

     3.    Subsection 3,  Field Operations,  provides  detailed  procedures  for  filter
          installation and recovery,  sample handling,  and data documentation.   It
          also includes  procedures  for the field  flow-rate calibration  check.
          Complete  documentation of background  information during sampling  is one
          of  several  QA  activities  important to  future data  validation; particular-
          ly  important information  are any unusual  conditions that  existed  during
          sample collection.  Such  conditions should  be noted.

     4.    Subsection  4,  Filter Preparation and Analysis,  presents  important consid-
          erations  for the handling, .integrity,  identification, equilibration,  and
         weighing  of filters.  A high-quality microquartz filter  is  recommended.
          Subsection 4 also  briefly describes minimum  laboratory  quality control
          (QC) procedures.  "The analytical  balance must be calibrated annually   and
         the  filters must be equilibrated in a controlled environment.

    5.   Subsection 5, Calculations, Validations, and Reporting of  PM10 Data
         presents calculations for determining PM10 mass concentrations and mini-
         mum data validation requirements.  The final data  review and validation
         including standardized reporting procedures, are all important parts of'a
         QA program.  Independent checks  of the data  and calculations are required
         to ensure that the reported data are both accurate and precise.

    6.   Subsection 6,  Maintenance, recommends periodic maintenance schedules to
         ensure that the equipment is capable of performing as  specified   The
         objective of a routine maintenance program is to increase measurement
         system reliability.

-------
                                                            Section  No.: 2.11.0
                                                            Date: January  1990
                                                            Page: 7


7.   Subsection 7, Auditing Procedures, presents independent audit  activities
     that provide performance checks of flow-rate measurements and  data pro-
     cessing.  An analytical process evaluation and a system audit  checklist
     are also provided.  Independent audits evaluate data validity.

8.   Subsection 8, Assessment of Monitoring Data for Precision and  Accuracy
     describes the assessment procedures for determining the accuracy and
     precision of the data.  The precision check is performed by using collo-
     cated samplers.                                        .         y

9.   Subsection 9, Recommended Standards for Establishing Traceability  dis-
     CxSu<:Vthe traceability of calibration equipment to established standards
     of higher accuracy.  Such traceability is a prerequisite for obtainina
     accurate data.                                                        s

10..  Subsections  10  and 11  contain the  PM10 Reference Method and  pertinent
     references used to prepare this document.   Subsection 12 provides blank
     data forms for  the convenience of  the user.

-------

-------
                                                                 Section No.: 2.11.1
                                                                 Date: January 1990
                                                                 Page:. 1


 2.11.1   PROCUREMENT OF EQUIPMENT AND SUPPLIES

      The establishment of an ambient PM10 air monitoring network requires the pro-
 curement of  specialized equipment and supplies for field operations and subsequent
 filter  analysis.   Information in this section has been provided to assist the agen-
 cy  in selecting  the proper equipment.  Subsection ,1.1  presents minimum sampling
 equipment necessary to conduct field operations.   Recommended laboratory instrumen-
 tation  is presented in Subsection 1.2.

      In  addition  to field operations and  laboratory equipment,  a data-handling
 system  (including forms,  logs,  files,  and reporting procedures)  must be developed
 and  implemented.   Sample  blank data  sheets are presented in  Subsection 12.

      It  is recommended that  each agency establish minimum monitoring equipment
 requirements  and  budgetary limits before  the  procurement procedures are initiated.
 Upon  receipt  of the sampling equipment  and supplies, appropriate procurement checks
 should be conducted.to determine their  acceptability,  and whether they are  accepted
 or rejected should  be  recorded in a  procurement log.   Figure 1.1,  which is  an  exam-
 ple of such a log,  will serve as a permanent  record for  procurements and provide
 fiscal projections  for future programs.   It will  also  help to provide a continuity
 of equipment  and  supplies.   Table 1.1, at  the end of the subsection,  lists  the
 major equipment needed, how  it  should be  tested,  suggested acceptance limits,  and
 actions  to be taken  if  acceptance limits  are  not  met;

 1-1   Procurement  Prerequisites—Field Operations

 I'l.l  HV PM1Q Samplers -  An  individual sampler must meet  U.S. EPA  operational
 standards and  be  a model  designated  as a  reference  or  equivalent method.  A  com-
 plete listing  of  minimum  sampler requirements  (i.e., 40  CFR  50,  Appendix J)  is
 reproduced in  Subsection  10.  Those  HV PM10 samplers not  designated  as  reference or
 equivalent methods may  not be used for reporting  data  to  determine  attainment  of
 the National   Ambient Air Quality  Standards  (NAAQS)  for particulate  matter.   The
 cost  of HV PM10 samplers will vary by manufacturer  and the options  chosen (i.e.
 continuous flow recorder,  timer).

      An in-house  inventory of general maintenance supplies and replacement parts is
 recommended.   These  include various handtools;  faceplate, motor, and  filter  cas-
 sette gaskets; genera.l  all-purpose cleaner; penetrating oil;-distilled water-  Kim-
wipes (or equivalent);  soft brush; and cotton swabs.  Spare  parts for the sampler
may be supplied by the manufacturer/ or many of them may be  purchased locally.

 1-1-2  Calibration Equipment - Calibration activities require specialized equipment
that will not necessarily be used in routine monitoring.  At a minimum, the  follow-
 ing equipment is  required.

          A thermometer capable of accurately measuring ambient temperatures to the
          nearest ±1°C  and referenced to a National Institute of Standards and
          Technology (MIST) or an American Society for Testing Materials (ASTM)
          thermometer within ±2°C (NIST is the former National Bureau of Standards
          [NBSJ) .

-------
                                                              Section No.:  2.11.1
                                                              Date:  January 1990
                                                              Page:  2
     z
     Ul
    o
    u
     &JJ

     01
    V)
    O
    u
Ul
0)
cc
   • "5
    o
    tr
    0

    i
    UJ
    u
    VI
    
-------
                                                               Section  No.: 2.11.1
                                                               Date:  January 1990
                                                               Page:  3
           A barometer capable of accurately measuring barometric  pressure over a
           range of  500 to 800 mm Hg (66  to 106 kPa) to the nearest millimeter of Hg
           and referenced at least annually to a standard of known accuracy within
           ±5 mm Hg.  For laboratory measurements, a Fort in- type,  mercury-column
           barometer  is appropriate.  For field measurements,  a  portable, aneroid
           barometer  (e.g.,  a climber's, or engineer's altimeter) is appropriate.
            pFH««      s*andard  (e;9" toP'hat orifice, variable orifice,  or
           !n  n5 Si™* capable of accurately measuring the  operational flow rate  of
           an  HV PM10 sampler at actual  conditions.  The transfer standard calibra-
           ^°nw^T r°nSh^? mus* be  refere"ced annually and be within ±2 percent  of
           the NIST-traceable .primary standard..                          n^v-cnt

           Water or oil manometer(s). with a 0- to 400-mm H20 (0- to 16-in )  ranae
           ta±an™nTm Sca1e.div^°n of 2 mm (0.1  in.)/ The VFC sampler  calfbra-
           fonn K6oUlrn T'"™6- a ue£?nd oil or water "noneter with a 0- to
           J ™"?S ?2° (?~ t0 36"in' H20) range and with a minimum scale division  of
     *     ecorderrrmf,rrpnrtS   fKth^SaTp1er is ^^^ "ith a continuous-flow
          recorder), miscellaneous handtools, and 51 -mm (2-in.) duct tape.

lio^of^he^y'pm^J6'!'06 " A ^flow-check device is required for routine opera-
tion of the HV PMio sampler;  a calibrated orifice transfer standard is recommended.


tionshio andfhp6r^naMfer/tandarJ TUSt have an NIST-traceable calibration  rela-
1  24 m3/n,?S  ?hS ?!r h^i ^.accur?^!y measuring volumetric flows between 1.02 and
 n *2 percent of the N?^? lr* re^tlon!hlP must ^ referenced annually and  be with-
                                                           '
                          r*                                               w-
                                                     ThiS °-'fi« transfer stand-

 1.1.4  Audit Equipment - The equipment needed for auditing is  similar to the cali-
 ^"frsriiEni %^^x
 l'2  Procurement Prerequisites— Laboratory Operation';
ter type selected!  reSenS   aS1C Cnteria that muSt  be met   gardless of  he  fil-

-------
                                                                Section  No.:  2.11.1
                                                                Date:  January 1990
                                                                Page:  4


1.2.2  Filter Protection - Post-sampling particle loss and filter damage will  occur
if proper handling procedures are not followed.  Filter cassettes are  recommended
for sampling with most HV PM10 samplers.  These may be purchased through the  HV
PM10 sampler's manufacturer.  A sufficient number of cassettes must be purchased to
allow insertion and removal of the filters in the laboratory.  For storage of
exposed filters,  however, cassettes may prove to be expensive and unwieldy.

     For assurance of the integrity of the exposed filter during handling and stor-
age, some type of protective covering is required; a manila folder in a protective
envelope is recommended.  The folder and envelope should be of comparable size
(large enough to allow easy removal of the filter, yet small enough to prevent
excess movement within the envelope) and be sealed to preclude damage or loss of
particles during transportation to the analytical laboratory.  The folder can be
printed to serve as a data sheet for proper documentation during sampling.   A suf-
ficient number of protective envelopes must be available to provide protection for
the filters during transportation to and from the monitoring location and for stor-
age of the exposed filters for subsequent gravimetric or chemical  analysis.

1.2.3  Laboratory Equipment - The analytical  balance must be suitable for weighing
the type and size of HV PM10 filters used.   The range and sensitivity depend on
routine tare weights and expected loadings.  The balance must be calibrated at
installation and recalibrated at least once a year,  as specified by the manufac-
turer.

     Prior, to their weighing, filters must  be conditioned in an  environment where
the mean relative humidity (RH)  is between  20 and 45 percent and controlled within
±5 percent, and mean temperature is between 15'  and  30°C and controlled within
±3'C.  Temperature and RH readings must be  recorded  daily,  either manually  or by
hygrothermograph.  Among the options available to ensure compliance with  the refer-
ence nethod specifications are a sling psychrometer  and a calibrated precision
thermometer (capable of measuring temperatures over  a  range of 10°  to 30°C  [283  to
303 K] to the nearest ±1°C)  that has been checked against an  NIST  or ASTM thermome-
ter to within ±2*C.

     It is impossible to present a complete procurement package  that would  provide
for unexpected contingencies in  any monitoring network.   Each  agency must determine
the extent of its in-house inventory and the  items that should be  ordered before
sampling can begin.  The agency  must also be  prepared  to  order any  additional
equipment required over and above that outlined  in this  subsection.

-------
Section No.: 2.11.1
Date: January 1990
Page: 5







V)
UJ
1-4
Q.
^
V)
S
K
i
a.
UJ
u.
o
UJ
§
o
OL
a.
o
u.
»-4
3 j
_J
it
V)
^•1
CO
UJ

UJ
1
t-
a.
UJ

••^
UJ
3
*- r
'
,
•



i

49
1
+>
*
«*.
"~ *
C •
0
i 1 \
j
L
3
er
t_



_2
"I
u
• c
49
a
u
0



WX
u

JE
o

u
c
a
• u
u







I
a.
•—
3
o-
UJ
.






t.
1
111
49
u
5

c
TJ
C ->
— X
3 -O
0 C
« a.
c a
0 <
e Q*
o 10
*• ee.
a. S
 V

L
1 O
'.05
•o c
2 •£ (
• •

e • • o
49 e u j=
e co 49
}JSI

u e c
t "o "e —
e c •
— e o >
a u — .-
E C H 3
e « e tr
v) -o -o e



—
c
0
u
e c.

2 1
e 

M
C
§ .2
^x 49
**j •
1 «
C Q.
0 0
•^ M
•^ C >h


v> ^~ o
jo a 49
— a. •
— 3 L.
• II o
0 JO






«
1
«»•
.^
u
*'

e —
"e -~
c -> '>
— X -0
49 .- C
3-D —
O C
e jc
c a —
0 < *
Specif icati
40 CFR B0,
Compatible
samplers.

u
e L
49 0
._ "^
—
-O
* 49

C 3
i -o
L. C
•™ •
tr-o •
e « o
t. a c
I I'i
.1 3 M









t.
u.

i ^
— c
L L
0 49
m t. •
— on
• «•- t.
• 0 —
49 JC —
E e
c OB
49 C
o e —
0 > 49
•-» o u
0 1. 0
e a Q.
ti

«~
— U
•• ^
s- a.
J:
C II
— o
u -o
0 C
rs,
= 1
•< T»



0
CO


•°
•**
o
^
u
c
0
T3
0
0



a
c
-o
C II

J= •

L. L.
49 4?
"•





j^
U
?
o

• •
3
T?

1
t.
O.
49
"I
X
>s 0
U 49
•
L. C
Within accu
presented i
•
2
e
u
•
i.

i
l»

Z
49
c
• •
co-o
• I.
e
JC T)
0 C
• •
J= 49
O «

49 C- !
C 0 1
I '^9
.— 5
3 e
'I
>S 1.
1- 49
O II
49 C
t.
o -o
-0 C
-J











































-------

-------
                                                               Section No.:   2.11.2
                                                               Date:   January 1990
                                                               Page:   1
 2.11.2  CALIBRATION PROCEDURES
      Before a PM10 monitoring  program commences, it is essential  to properly cali
 brate all  sampling and  laboratory equipment.  Calibration is defined as  the rela-
 tionship between an instrumental output and the input of a known  reference stand-
 ard.   Data that  .are traceable  to common reference standards are more uniform in
 character and more readily comparable than data that are not traceable.  Because
 PM10  concentration standards are not available for determining  calibration rela-
 tionships,  individual components of the sampling method must be calibrated to
 ensure the integrity of -reported data.

    "Like  a TSP  sampler, an HV PM10 sampler is essentially a device that pulls a
 sample of  ambient  air through a filter during a measured time period  and collects
 particulate mass on the filter.  Thus, to establish ambient PM10  concentrations
 three independent  determinations are made:  air volume flow rate, sampling time
 and particulate  mass.  The objective of this subsection is to provide technically
 sound flow-rate  calibration procedures for both the MFC and VFC HV  PM10 samplers.

 Note:   Calibration  procedures for samplers that have been  designated as reference
 methods  will  also  be provided by the manufacturer  in the operation  or instruction
 manual.  These manufacturer-provided calibration procedures are likely to be more
 specific and  detailed than the more generic procedures  presented  herein.   Also
 minimum  calibration  requirements for the determination  of  particulate mass  and'
 sample- time  are  presented in a table at  the end of  this  subsection  (Table 2.1).

      Although it concerns TSP samplers rather than  HV  PM10  samplers, Reference 3
 provides useful  information concerning flow-rate-calibration  procedures that  is
 applicable to PM10  samplers.   Reference  4  provides  useful  information concerning
 positive-displacement,  standard volume meters  (e.g., RootsR Meters).

      The following aspects  of PM10 monitoring  are discussed in this subsection:

     2.1    A discussion of volumetric and  mass  flow measurements  and their applic
            ability in  a PM10 monitoring program.                                ,

     2.2    A certification  procedure  for an orifice flow-rate transfer standard.

     2.3    Basic calibration  procedure specific to an MFC HV PM10 sampler.

     2.4    Basic calibration  procedure specific to a VFC HV PM10  sampler.

     2.5    Calibration  frequency requirements.

           2'1'  aVh! en
-------
                                                                Section No.:   2.11.2
                                                                Date:   January 1990
                                                                Page:   2


 (but  do not mix the two units).   Take care to avoid calibrating a PM10.sampler
 using one set of units and then  performing sample calculations using  another set  of
 units.

 2.1   Discussion of Flow-Rate Measurement and General  Aspects
      of PM10 Sampler Calibration

      An HV PM10 sampler consists of two basic components:   a  specially designed
 jnlet and a flow-rate controlling system.   The particle size  discrimination  charac-
 teristics of both the impaction  and cyclonic type inlets depend critically on main-
 taining certain air velocities within the  inlet;  a change  in  velocity will result
'in a  change in the nominal  particle size collected.   For this reason, it  is  impera-
 tive  that the flow rate through  the inlet  be maintained at a  constant value  that  is
 as close as possible to the inlet's design flow rate.  The  design flow rate for a
 given sampler is specified in the sampler's instruction manual.  The  manual  may
 also  provide tolerance limits (or upper and lower limits)  within which the sampler
 flow  must be maintained.   If the tolerance is not specified by the manufacturer,  it
 should  be assumed to be ±10 percent.   For  example,  if  the  design flow rate is spec-
 ified as 1.13 m^/min with  no tolerance given,  the acceptable  flow-rate range would
 be 1.02 to 1.24 m3/min.

      As indicated above, the true or" actual  flow rate  through the sampler inlet
 must  be known and controlled to  ensure that only those particles nominally less
 than  10 /p-are being collected.   A common  source of error  in  a PM10 monitoring
 program is confusion of various  atr volume flow-rate measurement units.  Although
 the sampler's operational  flow rate must be monitored  in terms of actual  volume
 flow  rate units (Qa),  sampler flow rates must be corrected to standard volume flow
 rate  units (Qstd)  at EPA standard conditions of temperature and pressure  to  calcu-
 late  PH10 concentrations as required  by EPA.  Thus",  both Qa and Qstd  flow rates are
 used  for PM10 measurements.   Before calibration procedures are initiated, the oper-
 ating agency personnel  should review  the following  flow-rate  measurement  designa-
 tions:

      •     Qa;  Actual  volumetric air  flow  rates,  measured  and expressed at existing
           conditions  of temperature and pressure  and denoted  by Qa (Qactual).
           Typical  units are L/min and m3/min.   Inlet design flow rates are always  •
           given in  actual  volumetric  flow  rate units.

      •     Qstd:   Air flow  rates  that  have  been corrected to equivalent standard
           volume flow rates  at EPA standard  conditions  of  temperature and pressure
           (25 °C or 298 K  and 760 mm  Hg or 101 kPa)  and  denoted by Qstd (Qstand-
           ard).   Typical units are std.  L/min  and  std.  m3/min.   Standard  volume
           flow-rate units  are often used by  engineers  and  scientists  because they
           are equivalent to  mass  flow units.   Standard  volumes  (derived from stand-
           ard volume  flow  rates)  are  required  for the  calculation  of  PM10 mass
           concentration  (/*g/std.  m3)  in reporting PM10  measurements.

      These Qa and Qstd measurement  units must  not be confused or interchanged.  The
 flow  rate units  can be converted  as follows,  provided  the  existing temperature and
 pressure (or in  some  cases  the average  temperature  and  pressure over  a sampling
 period)  are known:

-------
                                                                 Section No.:  2.11.2
                                                                 Date:  January 1990
                                                                 Page:  3
                            Qstd = Qa(Pa/Pstd)(Tstd/Ta)                        (Eq< ^

                            Qstd = Qa(Pav/Pstd)(Tstd/Tav)                     (Eq. la)

                              Qa = Qstd(Pstd/Pa)(Ta/Tstd)                      (Eq< 2)
  where:
            =  standard  volume  flow  rate,  std.  m3/min
            =  actual  volume  flow  rate,  actual  m3/min
            =  ambient barometric  pressure, mm  Hg  (or kPa)
            =  cnA sfan
-------
                                                               Section No.:  2.11.2
                                                               Date:  January 1990
                                                               Page:  4

                                     •
set of fixed resistance plates  (e.g., a reference flow [ReF] device or a top-hat
orifice), and the other with an externally variable resistance valve.  The series
of plates normally provided by  the orifice manufacturer includes an 18-, 13-, 10-,
7-, and 5-hole plate.  Unfortunately, the 5-hole plate provides too low a flow rate
to be useful for HV PM10 calibration, and other plates may produce flow rates sub-
stantially outside the design flow-rate range of the commercially available HV PM10
inlets.  Agencies may opt to fabricate or procure a different series of resistance
plates that will provide more flow rates within the sampler's .design flow-rate
range or use the variable-resi'stance type orifice device.

2.2.1  Orifice Calibration Procedure -

     1.   Assemble the following equipment:

          •   Orifice transfer  standard (i.e., top-hat orifice, variable orifice,
              or ReF device) to be calibrated.

          •   Water or oil manometer with a 0- to 400-mm  (0- to 16-in.) range and
              minimum scale divisions of 2 mm (0.1 in.).  This manometer should be
              permanently associated with the orifice transfer standard.

          •   Variable voltage  transformer (or a set of resistance plates, if a
              variable voltage  transformer is not available).

         '*•   Calibrated positive displacement, standard  volume meter (such as a
              Roots'* Meter) traceable to the National Institute of Standards and
              Technology (NIST).

              Note:  As they are sold,  standard volume meters may not be traceable
              to NIST.  Traceability can be established directly through NIST or
              indirectly through the meter manufacturer's repair department.  Peri-
              odic recertification is not normally required under clean service
              conditions unless the meter has been damaged and must be repaired.
              In general, damage will be indicated by a substantial (e.g., 50-per-
              cent) increase in the pressure drop across  the meter.  The meter's .
              traceability certificate should contain a graph of the pressure drop
              as a function of  flow rate.  See Reference  4 for additional informa-
              tion concerning standard volume meters.

          •   High-volume air mover (e.g., a blower motor from a HV sampler).

          •   Accurate stopwatch.

          •   Mercury manometer, with a 0- to 200-mm (0-  to 8-in.) range and
              minimum scale divisions of 2 mm (0.1 in.).

          •   Thermometer, capable of accurately measuring temperatures over the
              range of 0 to 50  8C (273 to 323 K) to the nearest ±1 °C and referenc-
              ed.to an NIST or  American Society for Testing and Materials (ASTM)
              thermometer within ±2 °C at least annually.                  .

-------
                                                              Section No.:  2.11.2
                                                              Date:   January 1990
                                                              Page:   5
            Barometer,  capable of accurately measuring ambient  barometric  ores-
                    r     "    °f 500. to 800 mm Hg  (66 to  106  kPa)  to  Jhe'nl r-

                                                                     °f
                                                                   «»"«•"  in
  2'   Sir °n  -5* "rt^ication  worksheet the .standard volume meter's serial
       number; orifice transfer  standard's type,  model,  and serial numb'er- 5-
       person performing the  certification;  and the date.          numoer, tne

  3.   Observe the barometric pressure  and record it as  Pa.
8.
                                                        ,.
 O.   v.iici_*  Lndr  rnp  <;r amiar-ri v/ninina ,„«*_- faklp •?«• 1   i   j  j •





 7'   fl"d  (os'Jtd SnrtJKr;C1»fij51.nA£'!ll,™?-t"- !1»« «0 avoid
for-!?n9er than 30 s at a time with
        eceockehi,n            -


-------
                                                                            Section No.:   2.11.2
                                                                            Date:   January 1990
                                                                            Page:   6
ORIFICE TRANSFER STANDARD CERTIFICATION WORKSHEET
Date: 7/ZZ./ ^^ Roots meter S/N: 7Y6V3/ £ Ta: 2.2. l M/m}' Qa = IUH2O (Ta/Pa)l": - b! i1/m!
•NOTE For PM10 monitoring, a calibration curve corrected to standard conditions is optional


                Figure 2.1. Example Orifice Transfer Standard Certification Worksheet.

-------
                                                          Section No.:   2.11.2

                                                          Date:   January 1990

                                                          Page:   7





      7pm v   and mercury.^nometers by sliding their scales  so  that their

      zero hnes are even with  the bottom of the meniscuses.
                             variable resistance  valve t  ach eve 'the appro




10.   After  setting a flow rate,  allow the  system to run for at least  1  min  tn


           '"0--  •
                                                                         '

                                                  ^lume meter (AVol.)  using




                  AVol. = Final Volume -  Initial Volume               (Eq.  3)



14.   Correct- this volume to ambient  atmospheric pressure.



                         Va =. AVol. (Pa - AHg)/Pa                      (Eq>  4)
    where:
                       V°]Ume  at  ambient barometric pressure  m3


            a   ambiet"? Z  T^^ by the Standard ^™* ^ter, m3
                ambient barometric pressure during calibration, mm iig .(or



          AHg = differential pressure at inlet to volume meter, mm Hg (or

-------
                                                          Section No.:  2.11.2
                                                          Date:  January 1990
                                                          Page:  8


15.   Calculate the actual volumetric flow rate (m3/min).

                               Qa - Va/ATime                           (Eq. 5)
     where:
            Qa = actual volumetric flow rate through the orifice, m3/min
            \ya = artnai vnlump at ambient barometric oressure, nH
                 Va  s  actual  volume  at  ambient  barometric  pressure,
              ATime  s  elapsed time,  min.                           ••      -     •

     16.   Repeat  Steps-9  through  15  for at  least  four additional  flow rates  within
          the  approximate range of 0.9  to 1.3 m3/min (32 to 46  ft*/min).   At least
          five evenly  distributed different flow  rates are required,  and  at  least
          three flow rates must be in the specified inlet  flow-rate  interval [1.02
          to 1.24 m3/min  (36  to 44 ft3/nrin)J.   Better calibration precision  may  be
          obtained by  running additional  flow rates or repeating  the flow rates.

     17.   For  each flow,  compute  [(AH20)(Ta/Pa)]1/2, and plot these  values against
          the  corresponding values of Qa.  Draw the orifice transfer standard's
          certification curve. For  the model  [(AH2<))(Ta/Pa)]l/^  = m(Qa)  + b,  cal-
          culate  the linear least squares regression's slope (m), intercept  (b),
          and  correlation coefficient  (r) of the  certification  relationship.  Plot
          the  regression  line on  the same graph as the calibration data.   A  certi-
          fication graph  should be readable to  0.02 m3/min.
          v
     18.   If any  calibration  point does not fall  within ±2 percent of the line,
          rerun that point, recalculate,  and replot.

     19.   For subsequent use of the  orifice transfer standard,  calculate Qa from
          the calibration-relationship  as:

                    Qa(orifice) - {[(AH20)(Ta/Pa)]l/2 . b} {1/m}             (Eq. 6)

          where:
              Qa(orifice) = actual' volumetric flow rate as indicated by the orifice
                            transfer standard,  m3/min
                    AH20 = pressure drop across  the orifice, mm (or in.) H?0
                       Ta = ambient  temperature during use, K (K = °C + 273)
                       Pa = ambient  barometric  pressure during, use,  mm Hg (or kPa)
                        b = intercept of the orifice calibration relationship
                        m = slope of the orifice calibration relationship.

2.2.2  Orifice Transfer Standard  Calibration Frequency -

     Upon receipt and  at 1-yr intervals, the calibration of the orifice transfer
Standard should be certified with a  standard volume meter  (such as a RootsR Meter)
traceable to NIST.  An orifice transfer standard should be visually  inspected for
signs of damage before each use,  and should be recalibrated if the. inspection re-
veals any nicks or dents.

-------
                                                              Section  No.:  2.11.2
                                                              Date:  January 1990
                                                              Page:  9
 2-3  Basic  Calibration Procedure for a Mass-Flow-Controlled
   .   (MFC)  Sampler Using an Orifice  Transfer Standard	
      The MFC sampler calibration  procedure presented in this  subsection relates
 arpW3ptP^n^eh t0 the Pressure  in  the exit orifice plenum.   The known flow rates
 are determined by an orifice transfer standard that has been  certified according to
               >resented in Section ? 9 1   THQ ovi+ -,_<*,•„„ ..i	  .  ... 	  3
2'2'1'   The  ex1t orifice P^num is the  area
 wthn  hpmt.-              '''                 e  Pnum  s   e area
 within the motor housing (below  the motor unit) that contains the air flow just
    °            a"     "
                                                    exit  orifice.  It is recom-
 wter^r^iT^nnSiJ1 '"H"  ^""u PreSSure 6e measured  W1"th a 25,cm (10-in??
 ifl Ln H^- man°meter-   I*  1S further recommended that  each sampler should have
 HLi*r hone?!    nianometer, which can be conveniently mounted to the side of the
 ?heS have romn?;,h?ther  types  of Pressure measurement devices may be used provided
 they have comparable accuracy.  However,  the 4-in.  continuous pressure (flow)
        accurate ^ °1 ten,suPP1ied ^th HV PM10 samplers are generally not suffi-
        *nt    Th    ff      recommended  for quantitative sampler pressure or flow

         Thft^?he fl°W !?as aPPrpxinjately  constant and uninterrupted over the sarnie
         The flow recorder may be connected  in parallel with the manometer or  othpr
pressure measuring device, using a tee or »Y» tSbing connection  man°meter Or  other

     Not§;   Because 'flow recorders are still  in wide use for Quantitative  finw


iE1?^£iTrSSS%^^^
     For this MFC calibration procedure, the following conditions are assumed:
             samp
         The sampler flow rate  is measured by measuring  the exit orifice plenum
         pressure,  using  a water or oil manometer [or,  if necessary  a coEtfnuous-
         flow recording device  using square-root-scale chart paperj!   C0ntinuou$-
         r^tP n-f i  n   
-------
                                                               Section No.:  2.11.2
                                                               Date:  January 1990
                                                               Page:  10               g^
2.3.1  Calibration Equipment -
     1.   Orifice transfer standard with calibration traceable to NIST (see Subsec-
          tion 2.2).

          Note;  The predrilled resistance plates that are supplied with the ori-
          TTce transfer standard may have to be modified for the calibration of
          PM10 samplers.  The holes in the plates may have to be enlarged or addi-
          tional holes may have to be drilled to obtain flow rates within the
          acceptable range.  Alternately, new resistance plates could be purchased
          if they are available.

     2.   An associated water or oil manometer, with a 0- to 400-mrn (0- to 16-in.)
          range and a minimum scale division of 2 mm (0.1 in.).

     3.   A water or oil manometer, with a 0- to 200-mrn (0- to 8-in.) range and a
          minimum scale division of 2 mm (0.1 in.) for measurement of the sampler
          exit orifice plenum pressure.  This manometer should be associated with
          the sampler.

          Note:  Manometers used for field calibration may be subject to damage or
          malfunction and should thus be checked frequently.

     4.  'Thermometer, capable of accurately measuring temperature over the range
          of 0 to 50 °C (273 to 323 K) to the nearest ±1 °C and referenced to an
          NIST or ASTM thermometer within ±2 °C at least annually.

     5.   A portable aneroid barometer (e.g., a climber's or engineer's altimeter)
          capable of accurately measuring ambient barometric pressure over the
          range of 500 to 800 mm Hg (66 to 106 kPa) to the nearest mm Hg and refer-
          enced within ±5 mm Hg of a barometer of known accuracy at least annually.

     6.   Miscellaneous handtools, calibration data sheets or station log book, and
          51-mm (2-in.) duct tape.

2.3.2  Multipoint Flow-Rate Calibration Procedure - MFC Sampler -

     The procedure presented here is basic and generic, given the assumptions list-'
ed in Subsection 2.3.  There may be more detailed calibration procedures, varia-
tions, or alternative calibration procedures presented in the manufacturer's in-  .
struction manual^.  It is recommended that the manual be reviewed carefully-and that
the various calibration variations or alternative procedures be evaluated.  In-
house equipment and personnel, procedural simplicity and uniformity, and subsequent
data applications should be considered in establishing the specific, detailed cali-
bration procedure to be implemented.

     Caution;  Do not attempt to calibrate the MFC sampler under windy conditions.
Short-term wind velocity fluctuations will produce variable pressure readings by
the orifice transfer standard's manometer.  The calibration will be less precise
because of the pressure variations.         -

-------
3.
                                                       Section  No.:  2.11.2
                                                       Date:  January 1990-
                                                       Page:  11
    • 2*2 iflllj^ji^ri^Vr*? " reSommended  fay the manufacturer.  Figure
   -  i.2 illustrates th

     stalledP

2.                                                                  into a
     22  il                                                      .      u
     i.2  illustrates the  calibration configuration of a typical MFC samoler
           P     3re Calibrated without  a  filter or filte? cassette in-
     Disconnect the motor from the flow  controller and plug  it directly i
     stable line voltage source (.i.e., the sampler's on-off  timer  if so
     equipped,  or other source of the  line vojtage).

     Install the orifice transfer standard and its adapter faceolate on th*
     sampler.   Check all gaskets .and replace any questionable oKes     .

     Caution:   Tighten the faceplate nuts evenly on alternate corners  to
     perTyjlign and seat the gaskets.  The nuts should be onlj hand-t^
     because too much compression  can damage the sealing gasket.      ll9
    5f]?cl'th? f!r!t Ca1i5ration fl°w rate and  install the appropriate
    resistance plate or adjust the variable orifice valve.  At  least four
    flow, rates are required  to define the calibration relationship

                  fl°W r/teS Sh°uld be within  (°r near1y within) the accent

                  ^ •?"""•£ - ™ -- «« a^:T£
    fceple      re in Pla" and ^riflce  is not cross-threaded on the

'    La      ''    ]     ^    '" ™th  *  1ar9e-diameter  rubber stopper  wide
                                                              "  '
                          ^6 Samp1fr for.;°n9er than 30 s at a time with

-------
Section No.:   2.11.2
Date:  January 1990
Page:  12
                a.
                o-
                «
                _o
                S
                .a
                "5
                u
               "a.

                a
                m
                03

-------
                                                           Section No.:  2.11.2
                                                           Date:  January 1990
                                                           Page:  13


 6.   Inspect the connecting tubing of both manometers for crimps or cracks.
      Open the manometer valves (if present) and blow gently through the tub-
      ing, watching for the free flow of the fluid.

      Adjust the manometers' sliding scales so that their zero lines are at  the
      bottom of the meniscuses.  Connect the orifice transfer standard manome-
      ter to the orifice transfer standard.  Connect the sampler's exit orifice
      manometer [and the continuous-flow recorder,  if used]  to the exit orifice
      plenum port.   Ensure that one side of each manometer is open to atmos-
    -  pheric pressure.   Make sure that the tubing fits snugly on the pressure
      ports and on  the  manometer.

 7.   [If a continuous  flow recorder is to be used  quantitatively in lieu of a
      manometer,  record the site location,  sampler  S/N,  date,  and the opera-
      tor s initials on the blank  side of a clean recorder chart.  Make sure
      the chart has  a square-root  scale.   Open  the  front  door of the sampler
      and install the clean recorder chart.]

 8.   Read and  record the  following  parameters  on the  HV  PM10  data sheet.
      Figure 2.3  presents  an example calibration  data  sheet'for  the  MFC sampler
      (blank forms appear  in Subsection 12).

          Date,  location,  and operator's  signature.
          Sampler S/N and  model.              ,                .    •
          Ambient barometric pressure  (Pa), mm  Hg or kPa.
          Ambient temperature (Ta),  K  (K  =  °C + 273).
          Orifice S/N and  calibration  relationship.

      Note:   Consistency of  temperature and barometric pressure  units  is
      required.  It  is  recommended that all temperatures be expressed  in kelvin
      in  »   t + Ml).   It  is also recommended that all barometric  pressures be
      expressed in either mm Hg or kPa  (but do  not mix the two units)    Take
      care to avoid calibrating a PM1.0  sampler  using one set of units and then
      performing sample calculations using another set of units.

      Note:   Ideally, the temperature of the air in the exit orifice plenum
      should be measured because it will be somewhat higher than ambient tem-
     5Sra-rS*K H°!^er\an adec*uate approximation of this temperature may be
     obtained by adding 30 K to the ambient temperature.  This addition is
      incorporated in the calculations given in Subsection 2.3.3.

9.    Turn on the sampler and allow it to warm up to operating temperature (3  •
     !°A^!;.,I?en:,,r"land..r!"rd the orifice transfer standard's manome-
                                                             rs manome-
deflPrt nn*1^' A?2° £"• H20)' and the "^Ponding sampler's manometer
deflection, APex [or flow recorder chart reading, I]!
                         may ^ Partia11y l«*red over the orifice
      prtH>
orov ded)   !£rj M ^ VS 3 ^ Shield (if a Shield 1s not otherwise
provTded).  Use a block to provTde at least 2 in. of clearance at the
bottom for air flow and for the manometer tubing.    tiearance    tne

-------
                                                  Section No.:  2.11.2
                                                  Date:   January  1990
                                                  Page:   14
MFC SAMPLER CALIBRATION DATA SHEET

Station Loa
Sampler Me
Pa _76,
Orifice S/N
Orifice calib
Plate
Number
/
*•
Z-
3
<-f-
5
*tinn /ty /£.
vtel ^^
^fnmHg, 1
' mm Hg, T
-2L.
ration relations
Total AH2O
(In.)
?.3-7-_.7C/^,.. , , _ _,_.
s* * » / °C .,*?(•;• / K, ('seafional average Ta and Pa)
Orifice Calibration Date 7/2-2^/
1 ' ' f Sampler so* point (SSP) ^t- ^ ^—<

SSP » [Pa/CTa + 30)] [m(SFR) + b]2
or SSP « [Pa/(Ta + SO)]'* [m(SFR) + b] for flow
recorders
Figure 2.3.  Example MFC sampler calibration data sheet.

-------
                                                                Section No.:  2.11.2
                                                                Date:  January 1990
                                                                Page:  15
      10.  Install the other resistance plates or adjust the variable orifice value
           to obtain each of the other calibration flow rates and repeat Step 9 for
           each.  At least four calibration flow rates are required,  with at least
           three in- the acceptable flow-rate range (i.e.,  1.02 to 1.24 m3/min).

      11.  Plot the calibration data on a sheet of graph paper as specified in
          'Step 4 of the next subsection.

           Note,5 .The data should  be plotted in. the fie.ld  as  the  calibration is
           occurring, rather than  afterwards back  at  the laboratory.
                  Step  V?r any  data  that  are  questionable  on  the  plot.   Running
           additional calibration points  at differing  flow rates or repeating  the
           calibration  points  at  the same flow  rates is encouraged  to  improve  the
           precision of the  calibration.

      12.   Turn off the sampler and remove  the  orifice transfer standard.

      13.   Reconnect the sampler  motor to the flow controller.

      14.   Perform the  calibration calculations presented in the following subsec-

       -    KMh^J-6   f! 9enerated V111 be used to set the mass flow controller
           h*~H nn ?£       ] *?  3 V3lue that wil1 reSMlt in °Ptim-al volumetric flow
           Dased on the seasonal  average temperature and barometric pressure at the
          monitoring site.

2.3.3  Calibration Calculations  -
             jf^^ f11 the cal ibration data,  including the orifice calibration
             the sampler calibration data sheet,  [and,  if used  the flow recorder
chart, which should graphically display the various calibration flow rates]
rather^n I!?!6 Ca^u1ations should be done at the time of the calibration,
t^.n -f     ?•   •   ™1S aPProach wl11  allow additional  calibration points  to  be
taken if questions  arise about the data that have already been  obtained
     l'    rnrilnt^^ ?* °riVCe transfer standard  calibrati-on  relationship  is
          current and traceable  to an  acceptable  primary  standard.
                                                                  th...Hf1c.-c.H-


                     Qa(orifice)  =  {[4H20(Ta/Pa)]l/2  . b}  {1/m}            '  (Eq. 6)
          where:
             Qa(orifice)  = actual volumetric flow rate as indicated by the trans-
                           fer standard, orifice, nP/min

-------
                                                          Section No.:  2.11.2
                                                          Date:  January 1990
                                                          Page:  16


                AH^O = pressure drop across the orifice, mm  (or in.) H?0
                  Ta = ambient temperature during use, K (K  « °C + 273J
                  Pa - ambient barometric pressure during use, mm Hg (or kPa)
                   b - intercept of the orifice calibration  relationship
                   m - slope of the orifice calibration relationship.

3.   Calculate and record the quantity APext for each calibration point as:

                        APext = [APex(Ta+30)/Pa]!/2                    (Eq.. 7)

     where:         •
          APext s transformed manometer reading
           APex = sampler manometer reading, mm (or in.) H20
             Ta = ambient temperature, K (K = °C + 273)
             Pa = ambient barometric pressure, mm Hg  (or kPa).

     [If a continuous-flow recorder is used quantitatively,  calculate and
     record the quantity It as follows:

                           It = I[(Ta+30)/Pa]l/2                      (Eq. 7a)

         where:
              It = transformed flow recorder chart reading
               I ~ flow recorder chart reading, arbitrary units on square root
                   scale.]

     Note;  If recorder charts with linear scales are used,  substitute (1)1/2
     for I in Equation 7a.

4.   On a sheet of graph paper, plot the calculated Qa(orifice) flow rates on
     the x-axis vs. the transformed sampler manometer response, APext [or the
     transformed flow recorder reading, It] on the y-axis.

     Because the determination of the sampler's average operational flow rate
     (Qa) during a sample period depends on the ambient average temperature
     and pressure, use of a graphic plot of the calibration  relationship is
     not recommended for subsequent data reduct-ion.  This plot is used only to
     visually assess the calibration points to see if any should be rerun.

     Plot the regression line on the same graph paper as the  calibration data.
     For the regression model y = mx + b, let y = APext and  x = Qa(orifice-) so
     that the model is given by:

                        APext = m[Qa(orifice)] + b                     (Eq. 8)

     [For the flow recorder,"the model is:

                          It-= m[Qa(orifice)] + b                   (Eq.  8a)]

-------
                                                              Section No.:  2,11.2
                                                              Date:  January 1990
                                                              Page:  17
           Using a programmable calculator or  a calculation data  form, determine  the
            i? JlHre9reS51;!! Sl°Pe '-intercePt (b), and correlation coefficient
           (r) and record  them on the data sheet.  A five-point calibration should
           yield a regression equation with  a  correlation coefficient of r > 0 990
           with no point deviating more than ±0.04 m3/min from the value predicted

           S^eth2rEr?Le9IJI*1^- iP1°Mthe Cession line on .the sS£ grjph
           paper that  has  the individual  calibration points.
      5'   nl^^fr*  -amp1? Pfriods- the sampler's  average actual  operational
           us'nglquat?^ 9?        ^ *"" *"* Ca1ibration *^* ™*  Intercept*1


                        Qi = {[APe7(Tav+30)/Pav]1/2  -. b) {1/m}             (Eq. 9)
           where:
                   = the sampler's average actual .flow rate,  m3/min

                   = ^P^Vlo^'wl1 and final sampler manometer  readings,
                     (APexi  + APexf)/2, mm (or in.) H20
                   = |verage ambient temperature for the sample period, K (K = «C  +

               Pav = average ambient pressure  for the sample  period  mm Ha (or kPa)
                 b = intercept of the sampler  calibration relationshiT
                 m = slope of the sampler calibration relationship.             '•

           [For the flow recorder,         '  '  -      '


                         Qi = {T[(Tav+30)/PaV]1/2-.b} {1/m}               (Eq. 9a)

          where:


              • I  - average flow recorder  reading for the  sample period..]
2'3'4  Hass Flow Controller Adjustment Procedure -








                           '
seasonal aveVa/^e^^'^^ ^     V£s j??^  «% "»P«t to the
 !^ta? ^ir&*]                       ™      »^ "°^TII
    t           (*                                  »                   .
volumetric flow rate his  at ?s and Js          "aSS """ "te " the inlet desi9"

-------
                                                               Section No.:  2.11.2
                                                               Date:  January 1990
                                                               Page:  18


     Note;  The correct SFR thus may be different from day to day and may be some-
what higher or lower than the inlet design flow rate on any particular day.

     Set the mass flow controller as follows:

     1.   Determine the seasonal average temperature (Ts) and seasonal average
          pressure (Ps) at the site and record them on the calibration data sheet.
          (Determination of the number of "seasons," i.e., the number of different
          seasonal average temperatures needed for the year, is left to the
          discretion of the local agency.)

     2.   Calculate SFR and record on the calibration data sheet:

                             SFR = (1.13) (Ps/.Pa)(Ta/Ts)                    (Eq.  10)

          where:

                  SFR = set-point actual volumetric flow rate for adjustment of  the
                        mass .flow controller, based on seasonal average temperature
                        and average pressure at site, m^/min
                 1.13 = inlet design flow rate (as specified by the manufacturer),
                        nH/min
               Ps, Pa = seasonal average and current ambient barometric pressure at
                        the site, respectively, mm Hg (or kPa)
               Ts, Ta = seasonal average and current ambient temperature,  respec-
                        tively,  K (K = °C + 273).                       .   '

     3.   Calculate arid record on the sampler's calibration data sheet the sampler-
          set-point manometer reading [or flow recorder reading]  that corresponds
          to the SFR calculated in Step 2.

                          SSP = [Pa/(Ta + 30)][m(SFR) + b]2                (Eq.  11)

          where:

               SSP = sampler set-point manometer reading,  mm (or in.) H20
                Pa = ambient barometric pressure,  mm Hg (or kPa)
                Ta = ambient temperature,  K (K = °C + 273)
                 m = slope of the sampler's calibration relationship
               SFR = set-point flow rate from Equation 10, -m^/nrin
                 b = intercept of the sampler's calibration relationship.

          [For the flow recorder,

                         SSP = [m(SFR)  + b]  [Pa/(Ta+30)]l/2              (Eq.  lla)]

     4.   Visually check  to make sure the motor is connected to the mass flow con-
          troller and the manometer is properly connected.

     5.   Install  a clean filter (in  a filter cassette)  in the  sampler according to
         'the manufacturer's instructions.

-------
                                                                 Section No.:   2.11.2
                                                                 Date:   January 1990
                                                                 Page:   19


            [If the continuous  flow  recorder  is  used  quantitatively,  install a  clean
            chart  and verify that the  recorder is zeroed  (i.e.,  the  pen  rests on the
            innermost circle of the  chart).]

      6.    Turn on the sampler and  allow  it  to  warm  up to operating  temperature (3
            to 5 min).

      7.    Following the manufacturer's instructions, adjust the mass flow control-
            ler until the manometer  reading [or  flow  recorder response]  indicates th<=
            sampler.set point (SSP)  as calculated in  Step 3.

      8..   Verify..that the flow controller will  maintain this flow rate for at least
            10 mm.  Turn off the sampler.

      9.    The sampler can now be prepared for the next sample run day.

 2-4  Basic Calibration Procedure for a Volumetric-Flow-Controlled
      (VFC) Sampler Using an Orifice Device~	

      The VFC sampler calibration procedure presented in "this subsection relates
 known flow rates  (Qa,  as determined by an orifice transfer standard) to the ratio
 of the stagnation pressure to  the  ambient barometric pressure (Pi/Pa).   The stagna-
 tion pressure (PI)  is  the air  pressure inside the sampler in the area just  under
 ine Ti-ner.  VFC  samplers have a stagnation  pressure tap or port throuqh" which the
 stagnation pressure'can. be measured.   A VFC  sampler may  also have an exit orifice
 r!t»Wrn ?HmKt0r S1mil*r  *« th?s* in MFC samplers.   In this  case,  the sampler  flow
 rate could be measured and calibrated using  the exit orifice plenum pressure  as
 described  in Subsection  2.3.   However,  the use  of  the stagnation pressurl generally
 provides a more accurate indication of  sampler  flow rate.   Additionally,  a  cSnfinu-
 ous-flow recorder may.be connected  to the exit  orifice pressure tap  for nonquanti-

 samjle  perioT"            ^  fl°W  rate W3S  constant and  uninterrupted  over the


 innn  Jt  /n  reco!Fmeride-d that the  stagnation pressure  be measured  with  a  0- to
 manrWo ?"  ?+ •  2*'*?11 Or water manometer (to  avoid  the  hazards  of  a mercury
 manometer).   It is further recommended  that  each sampler  have its own dedicated
 Other Jvn;,WnfCnrLan     conveniently  mounted to the  side of  the  sampler housing.
 Utner types of pressure  measurement instruments, may  be used  provided thev have
 comparable  accuracy.  However,  the  4-in.  continuous  pressure  (T.I    flow) recorders
 often s-upplied with HV PM10 samplers  are  generally not sufficientW  accurate and
 are not  recommended for  quantitative  sampler pressure or flow memlasurements ..

 nror,Ihe VFC seer's flow control system is a choked-flow  venturi.  It must be
 precisely sized for a given average annual temperature and pressure because no
puarchasinqPrace cl   L^not^f *?„adjustfthe operational  flow'rate  Ih^refore, the
J^nif   3-??   y     "    °tlfy the man"facturer of  the operational  location of the
the monjtorinali?rmlv rpf?ratUre a"d Pressure between  the shipping address and
MFC samoler  it  wi?!  h. n      In f" ^ncorrect  operational flow rate!  As with the
MFC sampler, it  will  be necessary to determine  or estimate both the ambient

-------
                                                               Section No.:  2.11.2
                                                               Date:  January 1990
                                                               Page:  20


temperature and barometric pressure readings during the sample period for the
subsequent calculation of total sample volume  in standard volume units.

     For this VFC calibration procedure, the following conditions are assumed:

     •    The VFC sampler uses a choked-flow venturi to control the actual volumet-
          ric flow rate.

     •    The sampler flow rate is measured by measuring the stagnation pressure
          ratio, and the sampler is not equipped with a continuous flow recorder.

     •    The sampler inlet is designed to operate at a constant actual volumetric
          flow rate of 1.13 m^/min, and the acceptable flow-rate range is ±10 per-
          cent of this value.

     •    The transfer standard for the flow-rate calibration is an orifice device
          equipped with either a series-of resistance plates or an integral
          variable-resistance valve.  The pressure drop across the orifice is meas-
          ured by an associated water or oil manometer.

     •    The sampler will be calibrated in actual volumetric flow-rate units (Qa),
          and the orifice transfer standard is also calibrated in Qa,  as specified
          in Subsection 2.2;

2.4.1  Calibration Equipment -

     1.   Orifice transfer .standard with proper calibration traceable to NIST (see
          Subsection 2.2).

          Note;  The predrilled resistance plates that are supplied with the ori-
          fice transfer standard may have to be modified for the calibration of
          PM10 samplers.   The holes in the plates may have to-be enlarged or addi-
          tional holes may have to be drilled to obtain flow rates within the
          acceptable range.   Alternatively, new resistance plates could be pur-
          chased if they are available.

     2.   An associated water or oil manometer, with a 0- to 400-mm (0- to 16-in.)
          range and minimum scale divisions of 2 mm (0.1 in.) for measurement of
          transfer standard pressure.

     3.   An oil or water manometer, -with a 0- to 1000-mm (0- to 36-in.)  range and
          minimum scale divisions of 2 mm (0/1 in.)  or other pressure  measurement
          device for measurement of the sampler stagnation pressure.   Ideally, this
          manometer (or other pressure instrument)  should be associated with the
          sampler.

          Note: Manometers used for field calibration  may be subject  to damage or
          malfunction and should thus  be checked frequently.

     4.   Thermometer,  capable of accurately measuring temperature over the range
          of 0 to 50 °C (273 to 323 K)  to the nearest  ±1  °C  and  referenced to an
          NIST or ASTM thermometer within ±2 °C at  least annually.

-------
                                                              Section No.:  2.11.2
                                                              Date:  January 1990
                                                              Page:  21
     5.   A Portable  aneroid barometer (e.g.,  a climber's or engineer's altime-
          ter),  capable of  accurately measuring ambient  barometric pressure over

                ±H°f >£°  t0c8°° r HQ (65 t0  106 kPa)  to the near*st ™ Hg and
                              mm  9 t0 3 barometer of known accuracy at least
                                                   b°°k and S1'm (2-in.)-wide

     7.    A clean  filter.

2'4-2  Mu^Point  Flow-Rate Calibration  Procedure - VFC Sampler -       '
                  , or alternative  calibration procedures    sted  in  the
                                  iS  rec°^ndedP that
                    -procedure  to be implemented
        blower motors  that are intended for use in  HV TSP samplers







                                         reSommended by the manufacturer.   Figure
                     n
         Step 8               '     ler may  e aPPr°P^'ate,  as  discussed  in

                            -----   -  -^
                  only,  too  much compression can damage the sealing gasket.

-------
                                                                              Section No.:   2.11.1
                                                                              Date:   January 1990
                                                                              Page:   22
(MOO mm H2O
Manometer
(0-16 in.)
                Resistance Plates
            18    13    10     7
0-1000 mm H2O
Manometer
{0-36 in.)
       APstg
       JL
                                      Stagnation
                                       Pressure
                                         Port
Calibration Orifice
Orifice Adaptor Plate
Rlter Paper
Cartridge/Cassette
                       Figure 2.4. Calibration assembly of the VFC sampler.

-------
                                                           Section  No.:  2.11.2
                                                           Date:  January 1990
                                                           Page:  23


 3.   Select a calibration flow rate and  install the appropriate resistance
      plate  (or no plate) or adjust the variable resistance valve.  At least
      four flow rates are required to define the calibration relationship.  At
      least three flow rates should be within the acceptable flow-rate range
      (i.e., 1.02 to 1.24 nP/min) for the sampler inlet.  For resistance plate
      orifices, make sure the orifice and resistance plate gaskets are in place
      and the orifice is not cross-threaded on the faceplate.

 4.   .Leak Test:  Block the orifice with a large-diameter rubber stopper  wide
      duct tape, or other suitable means.  Seal  both orifice and stagnation
      pressure ports with rubber caps or similar devices.  Turn on the sampler.

      Cautjon:   Avoid running the sampler for longer than 30 s  at a  time with
      the orifice blocked.   This precaution will  reduce the chance that the
      motor will  be overheated due to the lack of cooling air.   Such overheat-
      ing can shorten the motor's lifetime.   It  can  raise temperatures to the
      point of  defeating  the  electrical  insulation,  which could result in fire
      or electric  shock  to  the -user.

      Gently rock  the orifice transfer  standard  and  listen for  a whistling
      sound that would  indicate  a leak  in  the system.   Leaks  are usually  caused
        H e^ La  ^ama9?d or  ^ssing  gasket  between  the orifice  transfer  stand-
      ^H nn*I  %    ? ?te or bycrossthreading  of  the orifice  transfer  stand-
       •^ ^    taueplate'   A11  leaks must  be eliminated before proceeding
      with the  calibration.   When  the system is determined to be leak-free
      turn, off  the sampler and unblock the orifice.

      Note:   The leak test procedure that  is  described  above  is  one  of many
      alternate procedures which may be used.  Operating  agencies may develop-
      their own procedures.                                                 p

5.    Aspect the  connecting  tubing of the manometers for  crimps or cracks
      Open  the manometer  valves  (if present)  and blow gently through the tub-
   •   ing, watching for the free flow of the  fluid:

      bottnm ^e.J|anometers' sliding scales so that their  zero lines are at the
      transfer sSndW 5TJS'     fCt the transfer standard manometer to the
      ;™c    standard and the sampler stagnation- pressure manometer (or other
      ?fde of Mrh    Snt! t0-the Sta9nation Pressure port.  Ensure that one
      tubina fifJ '« T    *JS °Pen t0 atmc?5Pheric Pressure.  Make sure the.
      tubing fits snugly on the pressure ports and on the manometers.

6"          n   eC°rd te followin? Parameters on the VFC Sampler Data Sheet.
          Date,  location,  and operator's signature
          Sampler S/N and  model.

-------
                                                  Section No.:  2.11.2
                                                  Date:  January  1990
                                                  Page:  24
VFC SAMPLER CALIBRATION DATA SHEET
Station Location Gr-t-
•
\
(
I
Sampler M
Pa 7^
Orifice S/N
Orifice Cat
Plate
,No.
Notf£\
/^
'3
/&



odel W6.
— • « / ^^ ^% ^^ ^^ ^^ ^j ^ ™^i III ^r^^ ^*» ^^ ' ^?** .^^^\»^
*=. A/J' O K.Ti , v_ rl Date v / / / « / Time J~-' -3O " *rj
J»D/A^
~ S~mm Ha. Ta *i->. 7^

bration Rela
AH2O
(In.)
^75-
^./
^•f
^-5

"

Operational Row Rate
^«
tionship: m « _

APstg
(mm Hg)a
^3.3t/
*/7-,P£
^J.Vfi
&o.-) 
-------
                                                           Section  No.:  2.11.2
                                                           Date:  January  1990
                                                           Page:  25


      •    Ambient barometric pressure  (Pa), mm Hg  (or kPa) .
           Ambient temperature (Ta) , °C and K (K =  °C + 273).
           Orifice S/N and calibration  relationship.

      Note:  Consistency of temperature and barometric pressure units is
      required.  It is recommended that all temperatures be expressed in kelvin
      (K = °C + 273).  It is also recommended that all barometric pressures be
      expressed in either mm Hg or kPa  (but do not mix the two units).  Take
      care to avoid calibrating a PM10 sampler using one set of units and then
      performing sample calculations using another set of units.

 7.   Turn on the sampler and allow it to warm up to operating temperature (3
      to o in i n j •

      Note:  The sampler inlet may be partially lowered over the orifice
      transfer standard to act as a draft shield (if a shield is not otherwise
      provided).  Use a block to  provide at least 2 in.  of clearance at  the
      bottom for air flow and for the manometer tubing).

      Then, read and record  the orifice transfer standard's manometer reading
      AH20, and the corresponding sampler relative stagnation" pressure
      manometer reading,  APstg, on the  data sheet.   (Relative stagnation
      pressure is  a negative  pressure [i.e.,  a  vacuum]  relative to atmospheric
    .^pressure as  measured by a manometer with  one leg  open to 'the atmosphere.) '

      Note:  Be sure  to convert APstg to mm Hg  using Equation 12 before  record-
      ing  the reading- on  the  calibration data sheet:

                         mm Hg =  25.4(in.  H20/13.6)                    (Eq.  12)

8.    Install  the  other resistance plates  or adjust  the variable  orifice value
      to obtain  each of the other calibration flow  rates and  repeat  Step 7  for
      each.   At  least  four calibration  flow rates  are required,  with  at  least
      three in  the  acceptable flow-rate  range (i.e.,  1.02 to  1.24  m3/min).
             r    m*Lbe en"untered in obtaining flow rates in the acceptable
     range   Even with modified resistance plates (or with no plates)
     lr?th   5 UU may ^impossible to obtain three acceptable flow rates
     rll  a (i;te[.mounted on the sampler.. In this case, either lower flow-
     olatpS3 nSnaJh°n P°in? K?USt be USed and the ca1 ib™tion must be extrap-
     olated into the acceptable range,  or .additional calibration points must
            W1  °Ut * J] ] ter Jnst^ ^d in the sampler.  If additional calibra-
                  "6 °bta,;ned wlthout a ™ter, they should be examined care-
                   su£MheVre consistent with the calibration points
                         6r  le                                       all the
9.   Plot
     Step

-------

-------
                                                                Section No.:   2.11.2
                                                                Date:   January 1990
                                                                Page:   26


           Note:   The  data  should  be  plotted  in. the  field  as  the calibration  is
           occurring,  rather  than  afterwards  back  at  the laboratory.

           Repeat  Step  7  for  any data  that are  questionable on  the  plot.   Running
           additional  calibration  points  at differing  flow rates or repeating  the
           calibration  points  at the  same flow  rates  is encouraged  to  improve  the
           precision of the calibration.

           Turn off the sampler and remove the  orifice transfer standard.

     11.   Install a clean filter  on the sampler in the normal  sampling mode  (use a
           filter cassette if  one  is normally used).  Turn on the sampler  and  allow
           it to warm up to operating  temperature.

     12.   Read the relative stagnation pressure as in Step 7 and record it on the
           data sheet in the row for the operational flow  rate.

     13.   Perform the  calibration calculations presented  in the following subsec-
           tion.                .                                         3

2.4.3  Calibration Calculations -
      10,
e+a HGai*?er Aether all the calibration data, including the orifice transfer
standard -s calibration .information and the sampler calibration data sheet.
            Jhese calculations should be done at the time of the calibration
rather than later.  This approach will allow additional calibration poTnt  ?i be
taken if questions arise about the data that have already been obtained.
     1.


     2,
          Verify that the orifice transfer standard calibration relationship is
          current and traceable to an acceptable primary standard.

          Calculate and record Qa(orifice) for each calibration point from the
          orifice calibration information and Equation 6:

                     Qa(orifice)  = {[AH20(Ta/Pa)]l/2 . b} {1/m}        .     (Eq.  6)
          where:
               Qa(orifice)  = actual  volumetric flow rate as  indicated by the trans-
                             fer standard  orifice,  m3/min
                      AH20  = pressure drop across  the orifice,  mm (or in.)  H?0
                        Ta  = ambient temperature during  use,  K  (K = °C + 273)
                        Pa  = ambient barometric pressure during use,  mm Hg  (or kPa)
                         b  = intercept of  the  orifice transfer  standard's calibra-
                             tion  relationship
                         m  = slope of the  orifice  transfer standard's calibration
                             relationship.

-------

-------
                                                                Section No.:  2.11.2
                                                                Date:  January 1990
                                                                Page:  27


                                    PI  - Pa - APstg                         (Eq.  13)

           where:

                   PI  = absolute stagnation pressure,  mm Hg
                   Pa-= ambient  barometric  pressure, mm Hg
                APstg  = relative stagnation pressure,  mm Hg.

      4.    Calculate and record  the  stagnation  pressure ratio:

                        -   Stagnation pressure  ratio = Pi/Pa                 (Eq.  14)

      5.    On  a  sheet  of graph paper, plot  the  calculated orifice  transfer  stand-
           ard's flow  rates, Qa(orifice), on  the x-axis  vs. the corresponding  stag-
           nation  pressure  ratios, Pi/Pa, on  the y-axis.  Draw a smooth curve
           through the plotted data.  If necessary, extrapolate the curve to include
           the acceptable flow-rate  range.   '

      6.    If the  sampler manufacturer has  provided a  factory calibration table
       .    (i.e.,  the  lookup table)  for the sampler, compare Qa(orifice) for several
           points  on the calibration.plot with  Qa(sampler) determined from  the
           factory Calibration at Ta.  Calculate the percentage difference  between
           Qa(orifice)   and Qa(sampl.er) using  Equation  17:

             ± mff^r,^ , [Qa(sampler) - Qa(oriflce)] [1QQ]
                            IQa(orifice)J tlu°J          .     (Eq«-17)
                        i
           If the  agreement is within a few (i.e.,  3 or 4) percent, the factory
           calibration   is validated and may be used for subsequent sample periods
           Proceed to  Subsection 2.4.5.                  -

     7.    If the  agreement is not within a few percent, recheck the accuracy of the
           orifice transfer standard and recheck the calibration procedure.   Look
           for leaks, manometer reading  errors,  incorrect temperature or pressure
           data,  or miscalculations.   Also check for abnormally low line voltaqe at
           the site (it should be at least 110 VAC), for the  correct blower motor,
           and for the  presence of a gasket between the motor and the choked-flow
           ventun    A  factory calibration is not  likely to be substantially incor-
           rect,  and any discrepancy of  more than  a few percent is  probably  due to
           some problem with the  sampler or with the calibration procedure   How-
          ever  if no  errors  or  problems with the  sampler or with  the calibration
          can be found, or if  no factory calibration  is provided by the manufactur-
          er,  proceed  as described  in Subsection 2.4.4.

2-4-4  Generation  of Calibration Relationship - VFC Sampler  -

     1-   For each calibration point, calculate and record the quantity,

                                 Qa(orifice)/[Ta]l/2

-------

-------
                                                         Section No.:   2.11.2
                                                         Date:  January 1990
                                                         Page:  28
     where:
          Qa(orifice)  =  actual volumetric flow rate as indicated by the
                        transfer standard orifice, m3/min
                   Ta  =  ambient temperature during sampler calibration  K
                        (K = °C + 273)

2.   For the general linear regression model,  y = mx + b,  let  v  =  Pi/Pa and
     let.x = Qa(orifice)/[Ta]l/2, sucn that the model  is  given by:


                   Pi/Pa = m[Qa(orifice)]/[Ta] 1/2 + 5          .    '(Eq. 15)


                         re9ression s1ope ^ •  Intercept  (b) , and correlation
     the
                   'he plotted call'bratl'on  curve  to determine whether any of
                    points that are substantially outside of the acceptable
                   -need t0 be e!imi?ated so  tha' they do not cause an
                  linear regression line.
3.    For subsequent sample periods  the sampler's average  actual operating
                  , is calculated from the calibration  slope and intercept
                  n 16:                             •
                                                                      ng
    '   '      .   Qa(sampler)  =  {[(Pl./Pav).b][Tav]l/2}{.i/m}            (Eq." 16.)

    where:


         Qa(sampler)  =  the sampler's average actual flow rate,  m3/min
              Pl/Pav  =  average stagnation pressure ratio for the sampling
                       period                                       r    y
                 Tav  =  average ambient temperature for the sampling  period, K
                       \K ~  C •+• 273)
                   b  =  intercept of the sampler calibration relationship
                   m  =  slope of the sampler calibration relationship.

    Note:   The  average  value for PI should be calculated from staanation
    pressure  measurements taken before .and after the  sampliSJ peHod   Sav
    should be estimated from barometric pressure for  the sampling Seriod
    See also  Subsection 3.3 for additional information    Samplin9 peno-d'


    !I'Hn^1lbrati°n  (lookVp) table is  desired,'evaluate Equation 16 for
    various aoDrnnn^to u^inae  «-f 01/n, —i  T.  .  .  ,.  .   >-H«ai-'uu iu iui
    Note:  A calibration table based  on  Equation  16 may not match the
    ffl!^r--~ •?•

-------

-------
                                                                Section No.:  2.11.2
                                                                Date:  January 1990
                                                                Page:  29
2-4.5  Single-Point Operational Flow-Rate Verification -

     This procedure compares the VFC sampler's normal operating flow rate to the
desTgn flow rate of the inlet (e.g., 1.13 m3/min).

     1.    Determine the value of Pi/Pa for the operational  flow rate obtained with
          only the filter cassette installed (Steps 11 and  12 of Subsection  2.4.2)

     2.    Determine the sampler's operational  flow rate,  Qa(sampler)   that corre-
          sponds to this value of Pi/Pa.   Use  the manufacturer's calibration table
          if it has been validated in Step 6 of Subsection  2.4.3;  otherwise   use
          euation 16.
           equation 16.


                                                 '         rate
              Design  flow rate  %  difference  =  [Qa(sampler)  -  1.13jf100J      ^ lg)


           This design  flow  rate  percentage  difference must be  less than the'
           allowable  flow rate  tolerance  (i.e., ±10,  if~HoT otherwise specified by
           the manufacturer).   However, this value should be well within ±7  to allow
           for some variation with  ambient temperatUTe— If this value is not within
           ±7, recheck  the calibration procedure and  data for errors.  Check the
      "    5JS  «r Sh  1   n b?d m?t0r b™she*' missing gaskets, incorrect motor
           ld?«t!JiJ   ^ III  10W  I1"6 V°1ta9e-  Because the VFC flow rate  is not
           adjustable,  the VFC  manufacturer must be consulted to resolve cases of
           substantially  incorrect  VFC flow rates.                «u,ve cases OT

2-5  Sampler Calibration Frequency

c^nJ0 ensure.accurfte measurement of the PM10 concentrations, calibrate HV PM10
samplers upon installation and recalibrate as follows:          "MDrate Hv PMIU

.     1.   At least quarterly or annually according to the state's quality assurance

                   Se4R5AendiX A  ^ 3                   -""
                                       affeCt ^er calibration (e.g.,  replacing

     3.    After relocation of the sampler to a different site.

     4.    If the results  of a field flow-check exceed quality control  limits  (e  a
          greater than *7 percent from the sampler's  indicated  flow rate?!
     5.    Whenever a field flow-check  or performance  audit  indicates  that
                                             acceptabie

-------

-------
                                                     Section No.:  2.11.2
                                                     Date:  January 1990
                                                     Page:  30


Note:  Multipoint flow-rate calibrations should be distinguished from
single-point,  quality control flow checks (see Subsection 3.5).  The
latter are done more frequently than calibrations and are intended to
check if the sampler flow rate, Qa(sampler),  or the calibration relation-
ship have changed significantly since the last calibration.

-------

-------
                                                                                    Section  No.:    2.11.2

                                                                                    Date:   January 1990

                                                                                    Page:   31
C C    CM
o —  -  .
  •eJ A H4
  • u    c  I
  u c  c o
,-o0I —
  — u  = J    e
  — e — o    c
i  « *>.  o e    o
  o 0> c «
              fll "•• ^
             C      c
             O      o

             '•»>      4>
             o      o
             0 •»    0 m

             .g CM'    | «

             «/»       • -o  a>
          *> L    .- 0 ._
             e  x > *J  n
          •o *>  c a e  •
           e *-  • -a o T3

          •»   t> • -o  e

          3 J"  S -5.5 5

          e -2"  . *
          o 'e J "5 J? I
             u e   .»> v.
          -c • — »   *.
          •^ <- a. o e

          '* c » C £ 5
             O •   te* ««
          •      i.   Vi
          •» T3 C  O X
          • <- O   I- t.
          1- •    M «, o
          u> -o e *»      .
          — e  u — e  • e
          ~  • c -a  • j -u
          • •*  * 3 ji  • •
          W  «  C • *»  u "
          *)
1 1
1. •
31. —
a^e ?
*» — > u
« • — o
u «- c
-C
a u o o
•.S-J: i
» 0 • 1. -J
i — e ex
*> u i. > a
« e x o —
e a*» u «
•^ *
t» e «
x c • -> u
— — s x e
- ^ ..J-
u -o it e ~
— -x * %.
t TJ o
-* c e ts «.
f e x -u o
E u •
— e -p e
> u e c a
e • o • c
I- X — W •
« u •> • u.
•N.
J3 a
3-2-
x e .
1 U
•** e
e u
reading of w
chrometer on '
intervals.
-c >.
±ij|
.f"
U 3 4)
e ji e
a. i
6 x-o
0 L. C
o -o •
c
o
X
"0
4}
3
O
t.
•w •
c «
• X
e -o
e x
X L.
o e
u >
3 O
a o
•U L
e i
^
11
u •
o
u -u
0 e
Is
e 4
4> ^
•o3
i ;
•n e
c u.
•
•*> -k>
* e
e x «
u .—
5?;
»Si
U *J
?! • J
a c
^Si
0 C I

                    o
                                                        -t
                                                      - 1
                                                      UJ
                                       *ZJ
                                        S3?
                                       — O JJ

                                         "> II
                                       •O 0
                                        S E .2  *'

                                       5 e-s^
                                      .  = •• k  •
                                          e t.   ^
                                         *. e — e

                                           "
                                       * e  • i. >
                                       o u 4» o e
                                                                                   e
                                                                                   o

                                                                                   *j
                                                                                   u
                                                                                   0
                                                                                   n
                                                                                   J3
                                                                                   3   L.
                                                                                   U)   0



                                                                                   ^  1
                                                                                        U
                                                                                   •O    0
                                                                                                 I


                                                                                                O
 o

•O   3
_C   O

     C
 X   —
*>   ^>

•o   o.
.-   0

 3   e
                                                                                             0    *J


                                                                                             3    *
                                                                                             e   —
re
he
                                                                                                Q.

                                                                                               O

-------

-------
                                                                 Section  No.: 2.11.3
                                                                 Date: January  1990
                                                                 Page: 1
 2.11.3  FIELD OPERATIONS
      This section presents  information pertinent to the routine operation of a PM10
 monitoring site using an HV PM10 sampler.  It covers an array of topics, ranging
 from initial site selection to final data documentation.  The procedures herein are
 intended to serve as guidelines for the development of a monitoring programlhat
 will accurately reflect trends in local or regional air quality.  The effectiveness
 or the monitoring program depends on responsible day-to-day operation of the moni
     >nnnS^!"caTh? operatoCs who conduct sampling activities offer a unique perspec-
     -" data thSleL? Eth 0rmanCK'  ?nd their awareness and attention to detail-SiU
         fataJh:!^aL^!r!!l!Lbe 125t-  -Jt must be stress*d.  however, that 'You-
                                              operator provides  cohesiveness in  a
 3.1  Siting Requirements

      As with any type of air monitoring study in  which  sample data  are  used  to  draw
 conclusions about a general  population,  the validity  of the  conclusions  depends  on
 ±Uo?re-entatlTeSS °f ^sample data.   Therefore,  the pHrnary  goal
 monitoring project is to select  a  site or sites where the collected  par
 mass  is representative of the monitored  area.                         d
 ^mnil^b1^:1  Prese"ts  basic  S1'tin9  criteria  for the placement of the HV PM10
 sampler.   This  TS  not  a  complete  listing of  sitina requirements- instead  it s

 ^cation35  c'Lnf  fne-^ ^ °Peratl'ng  a^  t0 determinfa samp er'fopt mum
 ion   cn       -                                            erop
 }   < C°m? 6te S1tin? criteria are presented in 40 CFR 58, Appendix E   Addi-
            " °" °timUm ^"^                          '
 Reerence                                        site exposurecrtr   is given in
                        n0i s?eci!ied in the CFR must be considered in detenrnnina
                        e ^P10*6^  ™ese include accessibility under all weather
                         °f ad^ate.^^ricity, and security Jf the ^Jto^InT


     The sampler must be situated where the operator can reach it
   &sx
   i SeaudPfta f1?" <;•«•.  "librations,  filter installation  and  recovery,  flow
orin  s?     }           transporting supplies and  equipment to  and  from  the"
monitoing ste
sites with inrJX        Sa?1er it5elf  depends most1y on  its  location   Rooftop
sites with locked access  and  ground-level  sites with  fences  are  common.  In In
                                     personnel as well as  the  sampler should be

-------

-------
                                                                 Section No.: 2.11.3
                                                                 Date: January 1990
                                                                 Page: 2
                  TABLE 3.1.   MINIMUM HV PMl'O SAMPLER SITING CRITERIA
Scale
Height
above
ground,
meters
Distance from support-
ing structure, meters
Vertical Horizontal3
Other spacing criteria
Micro
Middle, neigh--
borhood, urban,
and regional
scale
                    2  to 7
                    2  to  15
>2       1. Should be >20 meters from
            trees.

>2       2. Distance from sampler to
            obstacle, such as build-
            ings, must be twice the
            height that the obstacle
            protrudes above the
            sampler.

         3. Must have unrestricted air
            flow 270 degrees around
            the sampler inlet.

         4. No furnace or incineration
            flues should be .nearby.b

         5. Spacing from roads varies
            with traffic (see 40 CFR
            58,  Appendix E).

         6. Sampler inlet is at least
            2  m but not  greater than 4
            m  from any collocated  PM10
            sampler.   (See 40 CFR  58,
            Appendix  A.)  -
    Un          or\ r00ft°?' tnis separation distance is in reference to
 walls, parapets, or penthouses located on the roof.

bDistance depends on the height of furnace or incineration flues, type of fuel or
 waste burned, and quality of fuel (sulfur, ash, or lead content .  This is to
 avoid undue influences from minor pollutant sources

-------

-------
                                                                   Section  No.:  2.11.3
                                                                   Date:  January  1990
                                                                   Page:  3
  3-2  Sampler Installation Procedures
                    J 'al'or«°ry "=!;«!; '» determine if the sampler fs operational
                        '                              motor  '              ™



                                                                    » instructions


      6:   Check'all tubing and power cords for crimps, cracks,  or breaks.
                                 of  inclement

     8.

     9.   Perform a multipoint flow-rate calibration, as described  in Section


3 - 3.  Sampl ing Operations
operation.  Significant difLrencerexi5t ?n th  J0^/""1"9 the samPler ^°

-------

-------
                                                                 Section  No.:  2.11.3
                                                                 Date:  January 1990
                                                                 Page:  4


      •    The  flow  rate  through  a  PM10  sampler  that  is equipped  with  a mass-flow
          controller  is  indicated  by  the exit orifice plenum  pressure.  This  pres-
          sure  is measured with  a  manometer  [or a  flow recorder].

      •    The  flow  rate  through  a  PM10  sampler  that  is equipped  with  a volumetric-
          flow  controller is  indicated  by the stagnation pressure.  This pressure
          is measured with a  manometer.

      •    The  sampler has been calibrated according  to procedures presented in
          Subsection 2.

      Sampling procedure  checks are summarized in Table 3.2 at the end  of this
section.

      The average actual  flow  rate  for MFC samplers is calculated by determining
(a) the average of  the initial and final manometer readings of the exit  orifice
plenum pressure [or the  average  flow recorder reading], (b) the  average  ambient
temperature (Tav)',  and (c) the average  ambient  barometric pressure (Pav)  during  the
sampling period.  These  values are then applied to the sampler's calibration  rela-
tionship.  The 4-in. pressure (flow) recorders  of the type often supplied with
HV PM10 samplers are generally not sufficiently accurate and are not  recommended
for quantitative sampler pressure  or flow rate  measurements.  These flow recorders
should be used only for  nonquantitative determination that the flow was  approxi-
mately constant and uninterrupted  over  the sampling  period.  The flow  recorder may
be connected in parallel with the  manometer or  other pressure measuring  device,
using a tee or "Y"  tubing connector.

      Note:  Because flow record'ers are  still in wide use for quantitative flow rate
measurements,  the procedures  in  this section include specific instructions for the
use of a flow-recorder.  These flow recorder instructions are enclosed in brackets


      The average actual  flow  rate  for VFC samplers is calculated by determining
(a) the average of  the initial and final relative stagnation pressures (APstg),
(b) the average ambient temperature (Tav),  and  (c) the average barometric pressure
(Pav)  during the sampling period and then by applying these values to  the calibra-
tion  relationship.

      Note:  Consistency of temperature and barometric pressure units  is  required.
It is  recommended that all temperatures be expressed in kelvin (K = °C + 273).   it
is also recommended that' all  barometric pressures be expressed in either mm Hg or
kPa (but. do not -mix the two units).  Take care to avoid, calibrating a  PM10 sampler
using  one set  of units and then performing sample calculations using another set of
units.

3'-3-!   Presamplinq Filter Preparation Procedures - Most HV PM10 samplers have been
designed to accept filter cassettes.   Loading these cassettes in the laboratory
will  minimize  damage;  however, if extreme care is exercised,  they can  be loaded at
the site when  ambient  conditions  permit.  Workers should wear protective gloves
when handling  filters  to avoid contaminating the filters with body oils and mois-
ture.   The filters should b.e  kept  in  protective folders or boxes.  Unexposed

-------

-------
                                                                  Section  No.:  2.11.3
                                                                  Date: January 1990
                                                                  Page: 5
wnen
               niter that has been labeled on its "down" side i


3-3-2  Sampling Procedures— MFC Sampler -
                                              at the



              of a cean

                                                        have
                        are  sed
          fro. th.,r curr,™"?^
                                                            '   If cllarts
                                                                        "

-------

-------
                                                           Section No.: 2.11.3
                                                           Date: January  1990
                                                           Page: 6


      [While  installing the chart, do not bend the pen arm beyond its  limits of
      travel.  Raise the pen head by pushing on the very top of the pen arm  (or
      by using the pen lift).  Be sure that the chart tab is centered  on the
      slotted drive to ensure full 360-degree rotation in 24 h.  Make  sure that
      the chart edges are properly located beneath the retainers.  Lower the
      pen arm and tap the recorder face lightly to make certain that the pen is
      free.]

      [Note;  During periods of inclement weather, the chart tends to  stick to
      the recorder face.  Two charts can be installed simultaneously to enable
      the sample (top, annotated) chart to rotate freely.]

5.    [Using a coin or slotted screwdriver, advance the chart and check to see
      that the pen rests on zero—the smallest circle diameter.  If necessary,
      adjust the zero set screw while gently tapping on the side of the flow
      recorder.  If a chart with a linear-function scale is used, some positive
      zero offset may be desirable to allow for normal variations in.the zero
      readings.]

6.    Turn on the sampler and allow it to equilibrate to operating temperature
      (3 to 5 min),

7.    While the sampler is equilibrating,  record the following parameters on
     the MFC Sampler Field Data Sheet (Figure 3.1):

          Site location.
          Sample date.
          Filter ID number.
          Sampler model  and S/N.
          Operator's initials.

8.    Inspect the manometer for crimps or cracks in" its connecting tubing.
     Open the valves and  blow gently through the tubing of the manometer while
     watching for the free flow of the fluid.   Adjust the manometer's sliding
      scale.so that its  zero line is  at the bottom of the meniscuses.

9.   Measure the initial  exit orifice plenum pressure (APex)  using an oil  or
     water manometer,  with a 0- to 200-mm (0-  to 8-in.)  range and a minimum  '
     scale division of  2  mm (0.1 in.).   Record  the initial  APex on the MFC
     Sampler Field Data  Sheet.   If APex  is substantially different than for
   •  previous samples  or  otherwise appears abnormal,  carry out a QC -flow check
     as described  in Subsection 3.5.1.

10.    [Verify that  the  flow recorder  (if  used)  is operational  and that the  pen
      is inking.   Note  the flow  recorder  reading.   If it  is substantially
     different than for  previous samples  or otherwise appears abnormal, carry"
     out a QC flow-check  as described in  Subsection 3.5.1.]

-------

-------
                                                                            Section  No.:  2.11.3
                                                                            Date: January 1990
                                                                            Page: 7
                              MFC SAMPLER FIELD DATA SHEET
 Station
 Location
                               fV c.
                                          Date
                                                                SAROAD*
Sampler Mode!
Filter ID No.
                       /M
                                                   S/N
                      2.?
                                        /
                                            . mm Hg, Tav
t &.
Final APex
        Sampler Manometer Readings                     Flow Recorder Readings
                            	in. H2O    Mean I	
                            	in. H2O
Mean APex
                     V. 0
                    */• /
                                    . in. H2O
Sampler Calibration Relationship:  m =   O ,

Q3  -  l-ld-2--
                                                        O ,
                                               Elapsed Time     /
 Qa «  {[meanAPexfTav + X)/Pavj'^ - b} {1/rn}
 Qa -  {mean I [fTav + 30)/Pavl* - b} {1/m} for flow recorders
 Operator
 Comments:  	
Laboratory Calculations:
Qstd =» Qa (Pav/Pstd) (Tstd/Tav)
Vstd
Vstd =. (Qstd) (elapsed time)
                               . std m3/min

                               	stdm3
                                             Gross weight (Wg)  _ "^ •  t
                                             Tare weight (Wt)       3>,<	
                                             Net Weight (Wn)      O .  d 4 & 7
                                             PM10 Concentration       3 9 . d
                                             PM10 Concentration « (Wn) (106)/Vtetd
                                                                                        K
                                                                            O , 'y^ 7 9
                                                                                      mjn
                         	Q
                         	'9
                         — g
                     . ^g/std m3
                     Figure 3.1. Example MFC sampler field data sheet.

-------

-------
                                                                Section No.: 2.11.3
                                                                Date: January 1990
                                                                Page: 8
     11.  Turn the sampler off.
     12.  [Check the time indicated by the time-set pointer on the flow recorder.
          If it i.s in error, rotate the chart clockwise by inserting a screwdriver
          or coin in the slotted drive in the center of the chart face until the
          correct time is indicated.]

     13.  Reset the elapsed time meter to 0000 min and the sampler timer for the
          next run day.  Close the sampler door, taking care not to crimp the
          vacuum tubing or any power cords.  The sampler is now ready to sample
          ambient air.                                         '            .    -

Filter Recovery Procedure - As soon as possible after sampling, the operator should
return to the monitoring site to retrieve the exposed filter.  Particle loss or
filter damage will result if the filter is left in the sampler for extended peri-
ods.           .

     1.   Turn on'the sampler and allow it to equilibrate to operating temperature
          (3 to 5 min).

     2.   Measure the final  APex and record it on the MFC Sampler Field Data Sheet.

     3.   Turn off the sampler.

     4.   [Open the door of the sampler,  remove the flow recorder chart,  and exam-.
          ine the recorder trace.  If the trace indicates extensive flow fluctua-
          tions,  investigate and correct  before the next sampling day.]

     5.   Record the .following parameters on the MFC Sampler Field Data Sheet:

          •     Elapsed time of the sampling period, min.
          •     Average recorder response, arbitrary units.
          •     Average .ambient temperature for the run day  (Tav),  K (K = °C + 273).
          •     Average ambient barometric pressure for the  run day (Pav),  mm Hg or
               kPa.

          Note;  The calculations presented in this subsection assume that the
          sampler has been calibrated in  terms o-f actual temperature and barometric
          pressure.  Average sampler flow rate for a sample period is determined
          from the average exit orifice plenum pressure (APex) and the average
          ambient temperature (Tav) and average ambient pressure  (Pav)  for the
          sample period.   Tav and Pav readings may be recorded or estimated on-site
          or may.be obtained from a nearby U.S. National Weather  Service Forecast
          Office or airport  weather station.   Barometric pressure readings obtained
          from remote sources must be at  station pressure (not corrected to sea
          level),  and they may have to be corrected for differences between the
          elevation of the monitoring site and that of the  airport.

-------

-------
4.
                                                              Section No.: 2.11.3
                                                              Date: January 1990
                                                              Page: 9


       fjote: Jf specific Tav or Pav values for the sample period cannot be
       tutPd"^; ?rS°nai avera9e temperature for the site (Ts) may be substi-
       tuted for Tav,  and average barometric pressure for the sit? fPO m*u L

       uX&'ZiS'*'  ^V^ "? t&^'  h°weve°' t^Y^L'  afcb  dV
       tions at  the site during the sample period can be reasonably reoresented
       by such averages    It is therefore recommended that seasonal valuef be
       used  to represent actual.values only within 20«C and 40 mm Hg!

  6'    .fln?1^6  an?.j;ecord  the .average actual  flow rate (as  determined bv  the
       ™dPner h  Call?rar°[! re!ati°nship)  on the MFC Sampler Fie fd Sat a Sheet
       and on the back of the  chart.   Attach the  chart to the data  sheet.


                {[APe7  (Tav + 30)/Pa]1/2  - b}  {1/m}                      (Eq- g)

       or for the flow recorder,


                9a)

      where


          -_2i = average sampler flow rate, actual. m3/min
          APex = average exit orifice plenum pressure, mm Hg or kPa'
            [I = average flow recorder response,  arbitrary Qnitsl
           Tav = average ambient temperature for  the run day  K
           Pav =  average ambient pressure for the run day, mm Ha or kPa
             b =  intercept of  the MFC sampler calibration relationship
             m =  slope of the  MFC samnler ral ihr-at,-«n ^i,*j'_.u7 5nip
                 ..-_.—r- -•  -•>- i'.v.  jamHici uaijoration  re  ationsi
            m = slope of the  MFC sampler calibration relationship!

     N£te:  If charts with  linear-function scales are used,  substitute
     (I) '* for T.


     Sayear??ect0ndfilterSDaart0?H ^ •?nit?Hn9 SUe;  "°te ^ activities'that
     mm-A Zu-    'ier Ro'ticie loading (e.g.,  paving,  mowino  fire^ and
     record.this information on the MFC Sampler Field Data Sheet   }


5'   -cilil?*n^l?r..!nl!* *nlle!!1?ve the f1Jter cassette.
6.   The sampler may now be readied for the next run day

7
  jamplinq Procedures--VFC Sampler -

-------

-------
                                                            Section  No.: 2.11.3
                                                            Date:  January  1990
                                                            Page:  10


 1.    Following the manufacturer's  instructions,  loosen the  nuts that  secure
      the  inlet to the base  and gently tjlt back  the  inlet to allow  access to
      the  filter  support  screen.

 2.    Examine the filter  support screen.  If the  screen appears dirty, wipe it
      clean.  If  the filter  cassette is equipped  with a protective cover,  re-
      move  it and place the  loaded  cassette in position on the sampler support
      screen.  Tighten the thumb nuts sufficiently to hold the filter cassette
      securely.  Check that  the gasket is in good condition  and has  not deteri-
   .   orated.

      Caution.-  Tighten the  thumb nuts evenly on  alternate corners to properly
      align and seat the  gasket.  The nuts should be only hand-tightened
      because too much compression  can damage the sealing gasket.

 3.    Lower the sample inlet and secure it to the sampler base.  For impaction
      inlets, inspect the sample inlet to make sure that it  is resting on the
      filter cassette and not on the PM10 sampler's frame.   Secure the sampler
      inlet to the sampler base.

 4.    Turn on the sampler and allow it to reach a stable operating temperature
      (3 to 5 min).                               •

 5.--   While the sampler is warming up,  record the following, parameters on the
      VFC Sampler Field Data Sheet  (Figure 3.2):

      •    Site location.
      •    Sample date.
          Filter ID number.
      •    Sampler model  and S/N.
      •    Operator's initials.                             .

6.    Bring an oil or water manometer to the side of the sampler.   This manom-
     eter should have a  range of  0 to  1000  mm (0 to 36 in.)  and  a minimum
      scale division of 2 mm (0.1  in.).

      Inspect the manometer for crimps  or cracks in its connecting tubing.
     Open the valves and blow gently through  the tubing of the manometer
     while watching for  the  free  flow  of the  fluid.

     Adjust the manometer's  sliding scale so  that its. zero line  is at  the
     bottom of  the meniscuses.

7.   Remove the vacuum cap from the stagnation  pressure port located on  the
     side of the sampler base.  Using  the connecting  tubing, attach  one  side
     of the manometer to the port.   Leave the  other  side  of  the  manometer  open
     to atmospheric  pressure.   Make sure  the  tubing  snugly fits  the  port  and
     the manometer.                       ,

-------

-------
                                                                          Section No.: 2.11.3
                                                                          Date:  January  1990
                                                                          Page:  11
                              VFC SAMPLER FIELD DATA SHEET
 Station
 Location
                          ,  CA
                                         Date
                                              5/7/37
Sampler Model    \jJ^DD)tiG-	

Filter ID No.  £77 J *•/   p-v   ISO
                                                              SAROAD*  OV./?/?V9
                                                  s/N
                                           mm Hg, Tav
                                                                          3O/-6
    Relative Stagnation Pressure Readings

Initial APstg

Final

Average
                       2	__ mm Hg    .Pi

                       L—_	mm Hg     Pi

                       	.	mm Hg

Average Stagnation Pressure Ratio (Pl/Pav)      O-

Average Flowrate (Qa)'   1-17*%     n_ ;i
•Obtained from manufacturer's lookup table (pr
 from alternate calibration relationship)
                                                      Absolute Stagnation Pressure
                                                       7ZO. 3
                                                                                  mm Hg
                                                  Pav -
                                                         Average .iPstg
                                              Elapsed Time
                                                                                    . mm
Laboratory Calculations:
Qsid
Qsta » Qa (Pav/Pstd) (TstdHav)
                               . Std m3/min
             /
Vstd
Vsta » (Qsid) (Elapsed Time)
                                   . std m3
                                            Gross Weight (Wg)

                                            Tare Weight (Wt)  _

                                            Net Weight (Wn)  _
                                                                 '.  3^02.
                                                                O. 07/7
                                            PMlO Concentration     55.
                                                                                    • 9
                                                                                    • 9
                                                                                     g
                                             PM10 Concentration m (Wn) (106)/Vstd
                                                                               wg/st£j
                    Figure 3.2. Example VFC sampler field data sheet.

-------

-------
                                                                Section No.: 2.11.3
                                                                Date: January 1990
                                                                Page: 12


     8.    Measure the initial relative stagnation pressure (APstg) and record this
          reading on the VFC Sampler Field Data Sheet.

          Note:  Be sure to convert the manometer reading to mm Hg using Equation
          12 before recording the reading on the VFC Sampler Field Data Sheet.

                            mm Hg = (25.4) (in.,H20/13.6)                   (Eq. 12)

     9.    Turn off the sampler, disconnect the manometer, and replace the vacuum
          cap on the stagnation pressure port.

     10.  Reset the elapsed-time meter to 0000 min and the sampler timer for the
          next run day.

     11.  The sampler is now ready to sample ambient air.

Filter Recovery Procedure - As soon as possible after sampling, the operator should
return to the monitoring site to retrieve the exposed filter.  Particle loss or
filter damage will result if the filter is left in the sampler for extended peri-
ods.-

     1.    Turn on the sampler and allow it to warm up to operating temperature (3
        --. to 5 min).

     2.    While the sampler is equilibrating, record the following parameters on
          the VFC Sampler Field Data Sheet:

          •    Elapsed time of the sampling period, min.
          •    Average ambient temperature for the run day (Tav), °C and K.
          •    Average ambient barometric pressure for the run day (Pav), mm Hg or
               kPa.

        •  Note:  Tav and Pav readings may be recorded or estimated on site  or may
          be obtained from a nearby U.S. National Weather Service Forecast  Office ,
          or airport weather station.  Barometric pressure readings obtained from
          remote sources must be at station pressure  (not corrected to sea  level),
          and they may have to be corrected for differences between the elevation
          of the monitoring site and that of the airport.  If Tav and Pav readings
          are not available, seasonal average temperature (Ts) and barometric pres-
          sure (Ps) can be substituted.  Care must be taken, however, that  the
          actual conditions at the site can be reasonably represented by such aver-
          ages.  It is therefore recommended that s'easonal values represent actual
          values within 20°C and 40 mm Hg.

     3.    Inspect the manometer for crimps or cracks  in  its connecting tubing.
          Open the valves and blow gently through the tubing of the manometer,
          while watching for the free flow of the fluid.  Adjust  the manometer
          sliding scale so that its zero  line is at the  bottom of the meniscuses..

-------

-------
                                                             Section  No.:  2.11.3
                                                             Date: January 1990
                                                             Page: 13


 4.   Remove the vacuum cap  from the stagnation pressure port  located on  the
      side of the sampler base.  Using the connecting tubing,  attach one  side
      of the manometer to the port.  Make sure that the tubing snugly fits the
      port and the manometer.  Leave the other side open to  atmospheric pres-


 5.   Record the final APstg on the VFC Sampler Field Data Sheet.  Turn off the
      sampler and replace the vacuum cap.

      Note:  Be sure to convert the manometer reading to mm  Hg using Equation
      12 before recording the reading on the data sheet.

                         mm Hg =25.4 (in.  H20/13,6)                    (Eq. 12)


 6.   Calculate the average relative stagnation pressure (APltg)  and record it
      on tne data she'et.

 7.   Calculate the average absolute stagnation pressure (Pi) for the sample
      run  day and record  it on the  data sheet.           -              ^mpie



 ,    "•   '                     P1  =  Pav  -  AFstg      -      .             .(Eq. 13)
      where   •  •  .      -                               .

              PI  =  average  absolute stagnation  pressure, mm Hg  or kPa.
             Hav  =  average  ambient  barometric pressure for the  run day  (not the
        ' 	    retrieval  day),  mm Hg  or kPa.                        	
          APstg  =  average  stagnation pressure  drop, mm  Hg or kPa.

8.    Calculate and record  the average  stagnation pressure ratio:
     where


          Pav


9.
                  Average  stagnation   •5T/D
                  pressure ratio     ~ P1/Kav                    (Eq. 14a)



      PI = average absolute stagnation .pressure, mm Hg or kPa
      aV = ane^9!Dambient barometn'c Pressure on the sample run day, mm
           • *yOiKiQ»
Using the  manufacturer's lookup table (or an alternate calibration
relationship as described in Subsection 2.4.4),  locale  the column and
row corresponding to Pl/Pav and_the  Tav value for the  sample run dav
Read and record the indicated Qa value.                 sampie run aay.

-------

-------
                                                                 Section No.: 2.11.3
                                                                 Date: January 1990
                                                                 Page: 14


      10.   Observe conditions around the monitoring site; note any activities that
           may affect filter particle loading (paving,  mowing, fire)  and record this
           information on the VFC Sampler Field Data Sheet.

      11.   Raise the sampler inlet and remove the filter cassette.  Replace the
           cassette protective cover (if so equipped).   To avoid particle loss,  be
        ,   care/ul to keep the cassette as level  as possible.

      12.   The sampler may now be readied for the next  sampling period.

      13.   Keeping the filter cassette level,  carefully transport it  and the data
           sheet to the laboratory sample custodian.

 3-3.4   Post-Sampling Filter Handling Procedures  .- If a sample will not  be  analyzed
 immediately,  the sample custodian should store the filter within a protective  cov-
 ering  to  minimize the loss  of volatile particles.   Because  filter cassettes often
 prove  too expensive and unwieldy for storage  purposes,  the  use of a  manila  folder  '
 and  a  protective envelope of comparable size  to  that of the filter is recommended
 Laboratory personnel  should adhere to the following  procedure:

     1.    Following the manufacturer's  instructions, remove the  top  frame of the
        .   filter cassette.

     2.    Conduct, a secondary  check  of  a sample's  validity  as  presented  in  "Labora-
           tory  Validation Criteria"  (Subsection  3.4).

     3.    Carefully slip  a  manila  folder underneath  the  edge  of  the  exposed filter.
           The filter  may  stick  in  the  cassette because  of overcompressi-on of the
           filter  cassette gasket.  Be  extremely  careful  to  avoid  damaqe  to  the
           brittle  quartz  filter.

     4.    Center  the  filter  on  the folder.  If the filter must be touched, do not
           touch or  jar  the deposit.   Fold the man.ila folder lengthwise at the  .
           middle with the exposed  side  of the filter in.  If the  collected sample
           is not centered on the filter  (i.e., the unexposed border  is not uniform
           around  the  f i Her) , fold it so  that only deposit touches deposit.  Do not
           crease the  folder—the sample  filter may tear;  If the  filter shears or
           breaks, ensure that all pieces  of the  filter are  included within the
           folder.

     5.    Insert the  folder  into the protective envelope.

     6.    Deliver the filter, in its protective folder and envelope,  accompanied by
           the completed data sheet, to the analytical laboratory.
              Analysis and Calculation of PM10 Concentrations - Post-samp 1 inq fil-
    analysis is discussed in Subsections 4.6 and 4.7.   The calculation of PM10
concentrations is discussed in Subsection 5.1.2.

-------

-------
                                                                Section No.: 2.11.3
                                                                Date: January 1990
                                                                Page: 15
     Sample Validation and Documentation -
                       -0^?* : The follow1'n9 criteria have been established to
 lefail  tn       h   Ini-*a1 y d!terminin9 wnetner a sample is valid.  If a sam-
Sbserved thJt ^  thes^cntena- do,nOt dl*scard the filter.  Document any factors
?n^   5*1  I   y resu1t in a Samp1e s invalidation oh the sample data sheet  and
t£p ?&? H data Sheet and the fl'lter to the laboratory supervisor  who wfll make
the final decision regarding the sample's validity.

     1.  - Timing:

               All  samplers must be turned ON and  OFF  within 1/2 h of midnight.

                                         ^ at  1east  ^ but ^™ than  2§ h
     2.    Flow  Rates:
             nd           ng period,  calculate  the percentage difference  between
            and  the design  flow  rate  (1.13 m3/min) using the following formuTaT
                          Difference. 100 'jZj                    .'    (£q>  lfl)
                  ^^lgate P°tential  err^ sources in^ediftely   T he followina
                  should be used as the basis for determining a sample's Validity?

              Decreases in flow rate during sampling (due to mechanical  nrohlpm^
              of more than 10 percent from the initial set poTnt result  in simp 1
                        !!!:;  «e"llbrat?  t«e sampler.   A saSple flow rate maJS so
                        due to heavy filter loading.   If a hiqh PM10 concentratinn
                                                  '           '
                                         Sh°Uld  inicate thisn      feld  data
              gardng                        ^  «"  ^ ^  ^  *^™  re-

-------

-------
                                                          Section  No.: 2.11.3
                                                          Date: January  1990
                                                          Page: 16
                             I
                                                       CO
  .1
                                                           o
                                                       o   2
  CM  ttj



     1
                                                               SI
                                                                CO
                                                                O
                                                                O
      o
  o
  CO
  o
  CM
                                                           vO
                                                              CO
tn

«3


_o


"o
C


15


15

•o
                                                                       r
                                                                       'CO
                                                                       J=
                                                                       U

                                                                       •"o
                                                                        o
                                                                        U

                                                                       o
                                                                       O

                                                                       .£
                                                                        a.
                                                                        E
                                                                        cs
                                                                        X
                                                                       HI
                  c

                  £
                  3
                                                       «   -i
                                                       Q   2
in
         o

         +•
                  m
in

-------

-------
                                                                 Section No.: 2.11.3
                                                                 Date: January 1990
                                                                .Page: 17
      1.   Check the filter for signs  of air leakage.   Leakage  may  result  from  a
           ?or*tP,r i?Pr?per1^ Insta11ed  faceplate  gasket.   A  gasket  general"? dlter-
            orates slowly.   The custodian should be  able to decide well  in  advance
           (by the increased fuzziness of the  sample outline) when to chanae t£
           gasket before total  gasket  failure  resets.  If  s  gns Sf  leakage are
      -     ^served,, void the sample,  determine the  cause,  and  instruct  th9e operator
           to take .corrective actions  before starting another sampling perfoS
           inorafternnn-         dama9e that «* have occurred dur
           da?e  [he  Hinif PJ  ?i  ?hysicai dama9e after sampling would not invali
           e I?             '0      f
is ,or?lqu,>ede.U5e °f '°9  6°0kS  "  PM1° ™°'t°'-™
-------

-------
                                                                Section  No.: 2.11.3
                                                                Date:  January  1990
                                                                Page:  18


     5.    Initial APex for MFC or  initial APstg for VFC  (DS)(LB).
     6.    Unusual conditions that  may affect the  results  (e.g., subjective evalua-
           tion of pollution that day, construction activity, weather conditions)
           (DS)(LB).
     7.    Operator's  initials  (DS).
     8.    Signature  (LB).

Operator Who Removes  the Samples -,                        '

     1.    Elapsed time of the sample run  (DS) (RC) (LB).   	.      _
     2.    Final APex  [or mean  I] for MFC or final APstg,  PI, and Pl/Pav  for VFC
           (DS)(LB)[RC].                              	
     3.    The calculated standard  average flow rate (Qstd) in std. m3/min (DS)(LB).
     4.   .The percentage difference between the actual and design flow rates (CC).
     .5.    Average ambient temperature and barometric pressure on the sample run day
           (DS)(LB).                                     •      ^
     6.    Seasonal average temperature and pressure, if needed (DS/LB).  This
           information needs to be  recorded in the logbook once, at the change of
           each season.
     7.    Existing conditions that may affect the results (DS)(LB).
     8.    Explanations for voided  or questionable samples (DS)(LB).
     9.    Operator's  initials (OS).
    10.    Signature  (LB).

3-5  Field QC Procedure - For. HV PM.10 samplers, a field-calfbratiori check of the
operational flow rate is recommended at least once per month.  The purpose of this
check is to track the sampler's calibration stability.  A control  chart  (e.g;,
Figure 3.4) presenting the percentage difference between a PM10 sampler's indicated
and- measured flow rates should be maintained.  This chart provides a quick refer-
ence of instrument flow-rate drift problems and is useful for'tracking the perform-
ance of the sampler.  Either the sampler log book or a data sheet  must be used to
document flow-check  information.  This information includes,  but is not  limited to
instrument and transfer standard model  and serial  numbers, ambient temperature and'
pressure conditions, and collected flow-check data.

     In this subsection,  the following is assumed:

           The flow rate through a PM10 sampler that'is equipped with a mass-flow
          controller is indicated by the exit orifice plenum pressure.  This pres-
          sure is measured with a manometer [or a flow recorder].

          The flow rate through a PM10 sampler that is equipped with a volumetric-
          •flow controller is indicated by the stagnation pressure.   This pressure
           is measured with a manometer.

          Both sampler models  are designed to operate at an actual  flow rate of
           1.13 m-Vmin. with an acceptable flow-rate fluctuation range of 10 percent
          of this value.

-------

-------
                                                                 Section  No.: 2.11.3

                                                                 Date:  January 1990

                                                                 Page:  19
                                                             CO
o   .2
                                                            o
                                                            CM
                                                                               en
                                                                              ^c
                                                                               u
                                                                               CD
                                                                              JZ
                                                                               o
                                                                              O
                                                                              O
                                                                              r
                                                                              a

                                                                              o
                                                                              c
                                                                              o
                                                                              u

                                                                              o
                                                                              O
                                                                              _0
                                                                              c.

                                                                              £
                                                                              en
                                                                              x
                                                                              UJ
                                                                              3
in

•t-
                                                            «
                                                            a
                                                                      0)
                                   133MO - DO

-------

-------
                                                                 Section  No.:  2.11.3
                                                                 Date:  January 1990
                                                                 Page:  20


      •    The  transfer  standard will be  an  orifice device  equipped with  a water  or
          oil  manometer.

      •    The  orifice transfer standard's calibration  relationship is  in terms of
          the  actual volumetric flow rate (Qa).

3.5.1  QC Flow-Check Procedure—MFC Sampler -
         ',      *  •   '
      The indicated flow rate  (Qa  (sampler)) for MFC samplers is  calculated by
determining  (a) the manometer reading of the exit orifice  plenum pressure [or the
flow  recorder  reading], (b) the ambient temperature (Ta),  and  (c) the  barometric
pressure (Pa)  during the flow check.  These values are then applied to the sam-
pler's calibration relationship.  The 4-in. pressure (flow) recorders  of the type
often supplied with HV PM10 samplers are generally not sufficiently accurate and
are not recommended for quantitative sampler pressure or flow measurements.  The
flow  recorder may be connected in parallel with the manometer'or other pressure
measuring device,- using a tee or  "Y" tubing connector.

      Note:  Do not attempt to conduct a flow check of PM10 samplers under windy
conditions.   Short-term wind velocity fluctuations will produce variable pressure
readings by the orifice transfer  standard's manometer.  The flow check will  be less
precise because of the pressure variations.

     An alternate QC flow-check procedure may be presented in the manufacturer's
instruction  manual:  It is recommended that  the manual  be reviewed and the various
methods be evaluated.   In-house equipment and procedural  simplicity should be con-
sidered in determining which method to use.
     1.
Collect the following equipment and transport it to the monitoring sta-
tion.  This equipment may be the same equipment as used for calibrations.

     A water or oil manometer with a 0- to 200-mm (0- to 8-in.) range- and
     a minimum scale division of 2 mm (0.1 in.), for measurement of the
     sampler's exit orifice plenum pressure.  This manometer should be'
     the same as is used routinely for sampler flow rate measurements.

     An orifice transfer standard and its calibration relationship.

     An associated water or oil manometer with a 0- to 400-mm (0- to 16-
     in.)  range and a minimum scale division of 2 mm (0.1 in.)  for
     measurement of the orifice transfer standard.

•   •  A thermometer capable of accurately measuring  temperature  over the
     range of 0° to 50°C (273 to 323 K)  to the nearest ±1'C and refer-
     enced to an NIST or ASTM thermometer within ±2°C at least  annually.

     A portable aneroid barometer (e.g.,  a climber's or engineer's altim-
     eter)  capable of accurately measuring ambient  pressure over the
     range of 500 to 800.mm Hg  (66  to 106 kPa)  to the nearest millimeter
     Hg and referenced within ±5 mm Hg of a barometer of known  accuracy
     at least annually.                                                J

-------

-------
                                                           Section No.: 2.11.3
                                                           Date: January 1990
                                                           Page: 21


      •    The sampler's calibration  information.

      •    Spare- recorder charts  and  a  clean flow-check filter.

           MFC Sampler Flow-Check  Data  Sheet (Figure 3.5) or site log book.

 2.   [Record the site location,  sampler S/N, and date on the back of a clean
      chart and install  it in  the  flow  recorder.  While installing the chart
      do not bend the pen arm  beyond  its limits of travel.  Raise the pen  head

                                           "*  "                  1?ft>
                                                                         that
 4'    t[h!tn?ha  coin,or slotted screwdriver,, advance the chart  and  check to see
      Il?i«5l  pen.hea? ^StS °" 2er° (*•«••• the smallest  diameter circle).  If
      necessary  adjust the zero-set .screw while gently tapping on the side of
      the  recorder.  A quarter turn of the set screw usually results in  arqe
      offsets;  adjust the set screw carefully.]             .               9

 5.    Set  up the  flow-check system as previously illustrated in Figure 2.2
      MFC  samplers are normally flow-checked  with  a filter  in  line (i e
   -  f?u!r-th*H°r1fiC? trar»fer standard and  the motor).  Instal  a clean
      f ufr lr     "if16";  Pla« thc  filter directly  upon the sampler's     -
      IlJ'iJhi    S'f    "K* USe a filter ca.ssette-   A  flow-check  filter should
      never be  used for subsequent s.ampling because particles  larger than  10 m
      can  be collected on the filter while the inlet'is  raised.  The sample  ^

                              result of  using a fiiter  for both • fi
6.   Install the orifice transfer  standard and its faceplate on the samoler
     Do not restr ct the flow rate through the orifice (e   bv usinq f xed
   •  resistance plates  or closing  the variable-resistance vaivej.      •

     HnrrSiirJl9!!1?"  th€ faceP1ate nuts on alternate corners first to elimi-
     HnT^fn  J   t0  unSUre even ^flhtening.  The nuts  should be hand-
     tightened;  too  much compression can damage the sealing gasket.   Make  sure
     the orifice transfer standard gasket is in place and the orifice transfer
     standard  is not cross-threaded on the faceplate.         oritice transfer

7.   Connect the orifice manometer to the pressure port of the orifice trans

                     s:-sss.r"         -^S1     L"

                                                                     r>    '
                                    connecti"9

-------

-------
                                                                            Section No.: 2.11.3
                                                                            Date:  January  1990
                                                                            Page:  22
                           MFC SAMPLER FLOW CHECK DATA SHEET
  Station     *j
  Location /W/LFOZQ,  flC	 Date //// /33      Time  /£>• 3Q
 Sampler Model    GMU	 S/N  _/2-T7____ Operator  D

 ^   7^3   mmHg. Ta 2T7. O   QQ  3OO.Q K< Unusual Conditions

 0rificeS/N     /O3<^7	Orifice Calibration Date

 Orifice Calibration Relationship:  m -   /. ZV3O   b , — O. OO6Z.  r m £). ef'J"	0/J  - bj  {1/mJ

  For calculation of sampler flow rates:

  Qa (sampler)  . {(QPex/Pa) (Ta +  30)1'^ - b} (i/m|
CQC o/o Difference  = [ °* (sampler) - Qa (or.fice) 1  r   -j
                   L       Qa (orifice)       J  I      w^ere Qa (sampler) is measured with the
                                                   J orifice installed

dQa (corrected sampler) = Qa (sampler) f 1°° " QC % ^wce 1
                               M   L          TOO                 e Qa (sarr>Pler) is measured
                                                          J
                                                                            r is mea
                                                             without the orifice installed.

eDes.gn Flow Rate o/0 Difference = I" Qa 
-------

-------
                                                            Section No.: 2.11.3
                                                            Date: January 1990
                                                            Page: 23


 8.    Turn on the sampler and allow it to warm up to operating temperature (3
      to. 5 min) .

      Note:  The  sampler, inlet may be partially lowered  over the orifice trans-
      fer standard to act as  a draft  shield  (if a shield is  not otherwise pro-
      vided).   Use a block to provide at  least 2 in.  of  clearance at the bottom
      for air flow and for the manometer  tubing.

 9.    Read and record the following parameters on the MFC  Sampler Flow-Check
      Data Sheet:

           Sampler location and date.
           Sampler model  and  S/N
           Ambient temperature (Ta) ,  °C and  K.
           Ambient barometric  pressure  (Pa), mm Hg or kPa.
           Unusual  weather conditions. :
           Orifice transfer standard  S/N  and calibration relationship.
           Operator's  signature.

 10.  Observe  the  AH20 across  the orifice by reading  the manometer deflection.
     Section  2.2  of  this  volume, .Figure 2.6,  illustrates the correct way to
     read  a water or oil  manometer.  Record the manometer deflection on the  •
     MFC  Sampler  Flow-Check Data Sheet.

 11.  Measure  the  exit orifice plenum pressure "(APex) by reading the manometer
     deflection-.    Record  the manometer deflection on the MFC Sampler Flow-
     Check Data Sheet.

 12.  [Using a coin or small screwdriver,  advance the recorder chart to read
     the sampler's corresponding response (I)  and record on the data sheet   A
     gentle tap on the recorder face  is often  necessary to ensure that the oen
     is not sticking to the chart.]
13.  Jurn off the sampler and remove the orifice transfer standard,  but not
     the filter.  Turn on the sampler and repeat Step 11 [or 12]  to  check the
     flow rate under normal  operating conditions.   Turn off the sampler and
     remove the filter.

14.  Calculate'and record Qa(orifice)  at actual  conditions  using  the following
     equation:    .                                                '          a
     where
     Qa(orifice) =.{[(AH20)(Ta/Pa)]1/2 . b} {1/m}        '    (Eq. .6)



Qa(orifice) = actual  volumetric flow rate as indicated by the ori-
              fice transfer standard,  m3/min
       AH20 = pressure drop across the orifice,  mm (or in.)  HoO
         Ta = ambient temperature, K
         Pa = ambient barometric pressure,  mm Hg or kPa
          b = intercept of the orifice calibration relationship
          m = slope  of the orifice calibration relationship

-------
                                                           Section No.: 2.11.3
                                                           Date: January 1990
                                                           Page: 24


15.  Calculate and record the corresponding sampler flow rate at actual condi-
     tions and record.

             Qa(sampler) = {[APex  (Ta + 30)/Pa]l/2 . b} {1/m}          (Eq. 9)

     or if a flow recorder is being used to measure the exit orifice plenum
     pressure,

               Qa(sampler) = {I[(Ta + 30)/Pa]l/2 - b} {1/m}            (Eq. 9a)

     where
          Qa(sampler) = sampler flow rate, actual m^/min
                 APex = exit orifice plenum pressure, mm (or in.)
                    I - recorder response, arbitrary units
                   Ta ~ ambient temperature during the flow check, K (K = °C +
                        273)'
                   Pa - ambient barometric pressure during the flow check, mm
                        Hg or kPa
                    b =- intercept of the MFC sampler calibration relationship
                    m = slope of the MFC sampler calibration relationship.

     Note;  If charts with linear-function scales are used, substitute (1)1/2
     for I.                                                      -  •

16.  Using this information and the formulas provided on the MFC Sampler Flow-
   •  Check Data Sheet, calculate the QC check percentage differences.
     QC-check % difference = [Qalsampler^-^orifice)]  [10Q]         (Eq>
     where Qa(sampler) is measured with the orifice transfer standard being
     installed.

     Record this value on the MFC Sampler Flow-Check Data Sheet and plot on
     the QC control chart.  If the sampler flow rate is within 93 to 107 per-
     cent (±7 percent difference) of the calculated Qa(orifice) flow rate (in
     actual volumetric units), the sampler calibration is acceptable.  If
     these limits are exceeded,  investigate and correct any malfunction.
     Recalibrate the sampler before sampling is resumed.  Differences exceed-
     ing ±10 percent may result  in the invalidation of all  data collected
     subsequent to the last calibration or valid flow check.  Before invali-
     dating any data, double-check, the orifice transfer standard's calibration
     and all calculations.

17.  Calculate the corrected sampler flow rate,  Qa(corr. sampler) usinq Equa-
     tion 23:

         Qa(corr. sampler) = [Qa(sampler)] [(10° " ^difference) j       (Eq>  23)

-------
                                                                 Section No.: 2:11.3
                                                                 Date: January 1990
                                                                 Page: 25


           where Qa(sampler) is measured without the orifice transfer standard being
           installed and where the QC-check percentage difference was obtained from
           Equation 17 above.

           Note;  Take care to use the correct sign (i.e., positive or negative)  for
           the -% difference when it is used in Equation 23.

      18.  Calculate and record on the MFC Sampler Flow-Check. Data Sheet the per-
           centage difference between the inlet's design flow rate (e.g., '1.13
           iiH/min) and the corrected sampler flow rate as:

           Design flow rate % difference = [Qa(corr-  sampler)  - 1.131  f10()J  (Eq>


           It is assumed in this subsection that the  inlet' is  designed to  operate at
           a  flow rate of 1.13 actual  nH/min.   If the  design flow rate percentage
           difference  is less than or equal  to ±7 percent,  the sampler calibration
           is acceptable.   If the difference is greater than ±7 percent,  investigate
           potential error sources and correct any malfunction.   Recalibrate the
           sampler before sampling is  resumed.  Differences exceeding  ±10  percent
           may result  in the invalidation  of all  data  collected  subsequent  to the
           last  calibration or valid  flow  check.   Before invalidating  any  data
           double-check the sampler's  calibration,  the' orifice transfer  standard's
           certification,  and all  calculations.

           Note:   Deviations from the. design  flow rate  may  be  caused in  part -by
           deviations  in  the site temperature  and pressure  from  the seasonal  average
           conditions.   Recalculate the  optimum set-point  flow rate (SFR) according
           to Equation  10  in Subsection  2.3  to determine  if the  flow controller
           should  be adjusted.

      19.   Set up  the  sampler for  the  next sampling period  according to the operat-
           ing procedure  in  Subsection 3.3.

3.5.2  'QC  Flow-Check  Procedure— VFC Sampler -             •

     The indicated flow rate  (Qa  (sampler)) for  VFC samplers  is calculated by
^rpTTTVaV?e^elKtive !ta9nation Pressure  (APstg) ,  (b) the ambient tempera-
ture  (Ta) • and  (c  the barometric pressure  (Pa)  during the flow check.  These val-
ues are then applied to the  sampler's calibration relationship.
            Dp not attempt to conduct a flow check, of PM10 .samplers under windy.
conditions.  Short-term wind velocity fluctuations will provide variable pressure
orPc^pShpr, <6 TtH" transfer standard's manometer.  The flow check will be less
precise because of the pressure variations.

     Note:.  An alternative QC flow-check procedure may be presented in the manufac-
turer's instruction manual.  It is recommended that the manual be reviewed and the
bfron^r^5 5H raluated' Jn-house equipment and procedural simp  cty  hould
be considered in determining which method to use.

                  the fonowing equiPment and transport it to the monitoring sta-

-------
                                                           Section  No.: 2.11.3
                                                           Date: January  1990
                                                           Page: 26
     •    A orifice transfer standard and  its calibration relationship in
          actual volumetric flow units  (Qa).

     •    An associated oil or water manometer, with a 0- to 400-mrn  (0- to
          16-in.) range and minimum scale  divisions of 2 mm (0.1 in.)-

     •    An oil or water manometer, with  a 0- to 1000-mm (0- to ,36-in.) range
          and minimum scale divisions of 2 mm (0.1 in.) or other pressure
          measurement device for measurement of the sampler stagnation
          pressure.  Ideally, this manometer (or other pressure measurement
          device) should be associated with the sampler.

          Note;  Manometers used for QC flow-checks may be subject to damage
          or malfunction and thus should be checked frequently.

     •    A thermometer capable of accurately measuring temperature over the
        .  range of 0° to 50°C (273 to 323  K) to the nearest ±1°C and refer-
          enced to an NIST or ASTM thermometer within 2°C at least annually.
          To calculate the orifice flow rates, it will be necessary to convert
          °C to K. '

     •    A portable aneroid barometer (e.g., a climber's or engineer's altim-
          eter) capable of accurately measuring ambient barometric pressure
          over the range of 500 to 800 mm Hg (66 to 106 kPa) to the nearest    •
          millimeter Hg and referenced within 5 mm Hg of a barometer of known
          accuracy at least annually.    .           .

     •    The sampler's calibration relationship (i.e., lookup table or alter-
          native calibration relationship).

     •    A clean flow-check filter loaded into a filter cassette.

     •    A VFC Sampler Flow-Check Data Sheet (Figure 3.6)  or a site log book.

2.   Set up the flow-check system, as previously illustrated in Figure 2.4.
     VFC samplers are normally flow-checked with a loaded filter cassette in
     line  (i.e., between the orifice transfer .standard and the motor).   The
     orifice transfer standard should be installed without fixed resistance
     plates or with  the adjustable resistance value fully open.

     A. flow-check filter should  never be used for subsequent sampling because
     particles- larger than 10 /*m can be collected on  the filter while the
     inlet is raised.   The sample mass will be biased as a result of using  a
     filter for both a flow check and subsequent sampling.

     Caution;   Tighten the faceplate nuts on alternate corners first to elimi-
     nate  leaks and  to ensure even  tightening.   The fittings should be  hand-
     tightened;  too  much compression can damage  the sealing  gasket.  Make sure
     the orifice gasket is in place and  the orifice transfer standard is  not
     cross-threaded  on the faceplate.

-------
                                                            Section No.: 2.11.3
                                                            Date: January 1990
                                                            Page: 27


 3.   Turn on the sampler and allow the sampler to warm up to operating temper-
      ature (3 to 5 min).

      Note:  The sampler inlet may be partially lowered over the orifice trans-
      fer standard to act as a draft shield (if a shield is not otherwise pro-
      vided).   Use a block to provide at least 2 in.  of clearance at the bottom
      for air flow and for the manometer tubing.

 4.   Read and record the following parameters on the VFC Sampler Flow-Check
      Data Sheet (Figure 3.6):

           Sampler location and date.
           Sampler S/N arid model.
           Ambient temperature (Ta),  °C and K.
           Ambient barometric pressure (Pa),  mm Hg or kPa.
           Unusual  weather conditions.
           Orifice transfer standard  S/N and  calibration relationship.
           Operator's signature.

 5.    Inspect  the  manometers  for crimps or  cracks  in  the connecting  tubing
      Open the valves and  blow gently  through  the  tubing, watching for  the free
      flow of  the  fluid.

      Adjust,the manometers'  sliding scales so  that the  zero  lines are  at the
   ...  bottom.of the meniscuses.         '                      -    -        •

 61    Connect  the orifice manometer to  the orifice transfer standard, and the
      sampler  manometer to  the  sampler  stagnation pressure port  located on the
      side  of  the sampler base.  Ensure  that one side of each manometer is open
      to atmospheric  pressure.  Be sure  that the connecting tubing snuqlv fits
      the  pressure ports and  the manometers.

 7.    Read  the pressure drop  as indicated by the orifice manometer (AH?0) and
      record on the VFC Sampler Flow-Check Data Sheet.  Read the stagnation
      pressure drop and record as APstg  (mm Hg) on the data sheet.

      Note:  Be sure  to convert APstg to mm Hg using Equation 12 before
      recording the reading on the data sheet.

                        mm Hg =. 25.4(in. H20/1.3.6)  '                  (Eq.  12)

8.   Turn off the sampler and remove the orifice transfer standard.

9.   With only a loaded filter cassette in  line, turn on the sampler and allow
     it to warm up to operating temperature.

10.  Read and  record the stagnation pressure drop (APstg)  for the normal  oper-
     ating flow rate.  Turn off the sampler.   Replace the vacuum cap on the
     stagnation pressure port.

-------
                                                                           Section  No.:  2.11.3
                                                                           Date: January 1990
                                                                           Page: 28
                          VFC SAMPLER FLOW CHECK DATA SHEET
I .ration  £\JEJJDQ2A,   CA	 Date f       m3/min       /• /^V      m3/min
QC — Check Percentage Differenced        ' • ^ _ •   _ q/0

Oa (corrected sampler)6 _ / • / / Cx _ m3/min

Design Row Rate Percentage Difference'     ~ '*  / __ 
-------
                                                            Section No.: 2.11.3
                                                            Date: January  1990
                                                            Page: 29


 11.  Calculate and record Qa(orifice) flow rate for the flow-check point  as
      in Equation (6) , reproduced below:                                 '

                Qa(orifice) - {[(AH20)(Ta/Pa)]l/2 - b} {1/m}            (Eq. 6)

      where

           Qa(orifice) = actual volumetric flow rate as indicated by the
                       .  transfer standard orifice,  m3/min
                  AH20 = pressure drop across  the orifice,  mm (or in.)  HpO
                    Ta-= ambient temperature,  K (K = °C + 273)
                    Pa = ambient barometric pressure,  mm Hg or kPa
                     b = intercept, of the orifice calibration  relationship
                     m = slope of the orifice  calibration relationship.

 12.   Calculate and  record the value of PI .(mm Hg)  for the  measurements  with
      and without  the orifice  installed.

                               PI  =  Pa -  APstg                        .(Eq.  13)

      where

              PI = stagnation  pressure, mm Hg  or  kPa
              Pa' = ambient  barometric  pressure, mm Hg or  kPa
          APstg = stagnation  pressure drop, mm Hg or "kPa.

 13'   w?!hU]nieJ?h/?C?£d  thVta9nation  pressure' ratio  for the measurements   "
      with and without the  orifice installed.

                     Stagnation pressure  ratio = Pi/Pa                (Eq. 14)

     where

          PI = stagnation pressure,  mm Hg or kPa.
          Pa = ambient barometric pressure, mm Hg or kPa.

14.  Refer to the instrument manufacturer's lookup table (or alternative cali-
   •  bration re atjonship as described in Subsection 2.4.4) and determine the
     Qa sampler) fow rates (n.3/min) for the measurements with and wUhout the
     ature In'"?   PP^ *H J"dicate?  for the r^io of Pi/Pa  and ambient  ?emPer-
  '   sJe                           U6S °n the  VFC samP1er fl°w check data
15.   Using Qa(orifice)  and Qa(sampler)  for the measurements with the orifice
     installed,  calculate the QC-check  percentage difference as:     oriTlce

     QC-check  %  difference =  [Qa(sampler)  - Qa(prificen  f    ]
                             I        Qa(orifice)        \  [100J         (Eq.  17)

     ?he°contthro?  chlrt  ?o 'nr JfC  ^ ]eJ  Flow-Check  °'ta Sheet  and  plot  it  on
     the  control  chart  for QC flow checks.   If  the QC-check  percentage  differ-

-------
                                                           Section No.: 2.11.3
                                                           Date: .January 1990
                                                           Page: 30


     ence is less than or equal to ±7 percent, the sampler calibration is
     acceptable.  Those differences exceeding ±7 percent will require recali-
     bration.  Differences exceeding slO percent may result in the invalida-
     tion of all data collected subsequent to the last calibration or valid
     flow check.  Before invalidating any data, double-check the sampler's
     calibration, the orifice transfer standard's certification, and all cali-
     brations.

16.  Using this percentage difference and Qa (sampler) from the measurements
     without the. orifice installed (i.e., for the normal operating flow rate),
     calculate the corrected sampler flow rate as:

        Qa (corr. sampler)  = [Qa (sampler)]  [10° " ^fferencej      (Eq> 23)


     Record Qa (corr. sampler)  on the VFC Sampler Flow-Check Data Sheet.

17.  Determine the design flow rate percentage difference between the inlet
     design flow rate (e.g.,  1.13 m3/min) and Qa (corr. sampler) as:

     Design flow rate % difference = [Qa (corr.^sampler) - 1.13J      (£q> lg)


     Record this design flow rate percentage difference o,n the VFC Sampler
    -Flow-Check Data Sheet  and  plot it on the control  chart for the field
     validation of flow rates.   When plotting this value,  use a different
     symbol  than is normally  used for plotting values  that are obtained during
     sampling periods.   If  the  design flow rate percentage difference is less
     than or equal  to ±7 percent,  the sampler calibration  is acceptable.
     Those differences  exceeding *7 percent  will  require recalibration.   Dif-
     ferences exceeding ±10 percent may result in the  invalidation of all  data
     obtained subsequent to the last calibration  or valid  flow check.  Before
     invalidating any data,  double-check the sampler's calibration,  the ori-
     fic'e transfer standard's certification,  and  all calculations.

-------
                                                                                       Section  No.:  2.11.3

                                                                                       Date:  January  1990

                                                                                       Page:  31
                                                       a
                                                      x
+3 • "•
1 , c ^















CO
u
HI
• 3.2. SAMPLING PROCEDURE CH
Z S
e 49
.'1 .!-.
3 ' U
e o IT 49
i. c • a —
••» A — — --
„ f i ~ *•

• •*— 9
J ' | I
0 49
"3 3

1 ' 3 « 49 3
0 E 3 — C 6
i- e * a •
a. a a 49 a
49 t c u a a
xi 49 c e u jc a
c a 49 a n ti ^
Requirements
v
Filters must be .undamaged a
vided with ID numbers; cess
be centered, on sempler; fil
be centered on cessette scr
edges peril lei to edges of
end to feceplete gesket; ge
be tightened to prevent lee
._
e a
x > •
49 a e -
c —
49 3 0.
U E
a s. •
c. — a
o — a
U C X
O 49
L ••
O 4> TI
a *'o
49 C >
-? I .
g-S-S
O O X
o-o a
i
a ••
X L. --D
49 o a c
— 49 O
* Q. a o
0 E L.
.f A ^
Racord on data sheet must si
sempl ing dete, filter ID, si
S/N, stetion iocetion, flow
semple time, and eny unusua
tions.
O — 49
— 3 a a
»•  49 _C
0 <- C U
— O >-t JC
49 U *
• 49 . a o
t c • x —
a — t. «•-
o. o a «
o a. u a a
« a "o i"S
3 a a a 49
**^ **• N &
•o a 3 a
a 4> c — u
a a a — u
a: u E a a

_ ^
i *
• =
a a
I. -C
49
o •»•
— o
«*•
K
— ®
a f->
c ^ -
o a -
•- O 49
49 49 a
a • L
• §> *
0. O
o c —
•— «*-
u x
e 49 c
*•;.?
E «
a e o
CO X- 73
49
u

L •
0§
u —
4.1
— 0
a c
cm-
E —
a a
• E
• «•-
X O
49
a
~ I
o a
> u
i.
.3

«». •
c
There must be no evidence of
damege or sampler melfunctic
 -o   —
 o   «^ c

.5 '  .cJ

 I    i";

 -o1   « .
 c
 a
      u a
      a -t
 x ;
 u    u a
 C ;     CO
 a    x
 3   — *>  .
 v   — u a
 a .   • a —
 <- •   3 a a
U.    M « E
     •- C •
     > H-l W
 e    i  c
 t- .   co
 3   .- ._
~D     49
 a    L.  •
 u    a —
 O '  49 —
                         ts
 a -o «
     M
jc e c
 o — c
 a a. a
-C 6 49
 u a a
  • •> —
 x   a.
— a E
— -co
 a 49 u
 3
 •I TI l_
                         I Tt
                        —  C I
                        a  a a
                        >    4>
                           c c

                        •.2 I
                        0.49 3 c

                        E  a o o
                        a TI o —
                        CO — T> 4>
 49 4>
 a e x
 C. —  u
   c  a

 5-  •
 — X  U
 •>• 49  O
   •— 4te
 — t
 a    a
 c a 49
 O u  a
•- c  u
 4> •
 a E  s
 «. i.  o
 a o —
 CL-~ «-
 O c
   oca
-X U Ol —
 o   —  a
 c t. n  E
x o a  •
                                                     a
                                                     « u
                                                       o
                                                    x
                                                     u n
                                                     a a
                                                     a u
                                                       a

                                                     u a.
                                                     a
                                                    X CO
                                                     u c

                                                     X «  •'
                                                    — «i a
                                                    — — ra
                                                     a E a
                                                     3   JC
                                                     * I. *
                                                    — o a
                            §•-2
                            a 49
                            «  o
                            i  e
                            ••9  Q.
                            «i  n
                            o  c
                            a. —
                                        «
                                        a         i
                                        u    xix
                                        c  i  a —  3
                                        « — x —  n
                                        i. — 49 a
                                        « a    > TI
                                       »- u «•- c  a
                                       *• a w — 49
                                       — i.       o


                                       1 J J if-

                                        M.S s s
                                       49 u a. u
                                        c u E e •
                                        • a a o.4>
                                        u a «   a
                                        L.     S "O

                                        §.7 * "

                                        a TI   TI "a
                                       x a a •
                                       *> 0 49 a •
                                          u a u 49
                                       o- x i. x a
                                       M a x a -o
                                                               e
                                                               o  a 49
                                                               — -o c
                                                               *»    a
                                                               — •DO
                                                               •O — L
                                                               e  3 e
                                                               o  o a
                                                               U X
                                                                  « !«.
                                                               C    ti
                                                               co  n
                                                               —  • o
                                                               «  u 49
                                                               o  c
                                                               -a  a —
                                                                  L. a
                                                               •e  a 3
                                                               c «.. 
                                                               « L.  ti
                                                              x a  a
 o

 4>  •
 • X
 U 49
 X C

 = i

 U I.


 S*
 a e
 i. u
 i  c
 » o
 o
— 49
jc  a
u —
                        w
                       -1C
                        u
                        a
                                                                            i
                                                                            I-   O X
                                                                           x-  • i. 4> a
                                                                           — je L   x
                                                                           — o a  • *>
                                                                            a e    L.
                                                                            o x —  a TI
                                                                              u a — c
                                                                           4>   —  a. a
                                                                            » * 4>  6
                                                                            a o e  a •
                                                                           	«  11 a
                                                                             *. 49   49
                                                                            a   oca
                                                                           x T> o.x -o  •
                                                                           49 •-   49   ii

                                                                            o a 4>  coo
                                                                           4) > a — — .-
                                                                                a   49 4>
                                                                           •a. u •-  n a a
                                                                            c o 49  a <- —
                                                                            •   « o -a 3
                                                                            3 c e i. — o
                                                                            »• o > s — —
                                                                            « — c o a  a
                                                                           « 49 M a u  u

-------

-------
                                                                 Section No.: 2.11.4

                                                                 Date: January '1990
                                                                 Page: 1
 2.11.4  FILTER PREPARATION AND ANALYSIS
                        ?M!° sai"P1in9 Program depends on several factors.  A primary

 naue   Th! labora*°^ staff's attention to detail and balance tech-*
 nique.  This section offers guidelines to enhance the accuracy of the laboratory
 operation and, hence, the determination of PM10 mass concentrations   laboratory



 PM10 fiUe?s:Cy ""^ ** aW3re °f two Pn'mary sources of error in the handling  of HV
      1.   Loss of particles during shipment  or handling.   Subsection  4  1  oresents
           guidelines to help prevent  post-sampling  particle  loss.         Presents.
           t^thf fn™™f 10?Arr°rS  a^e  ^aused  by  the  retention of sulfur dioxide
          •in the form of sulfate  particulate on  alkaline  filters.  The results of
           experiments involving a variety of filters indicated that sulflte loadina


                             '

                                       0? • Pr?«dulr« for *»" laboratory apparatus


                                             "''    '
                                     currently the ""'y cornnercfally available HV
4.1  Filter Handli
                  ng

                              i                    °
         to                        '

-------
 Criteria
                                                                Section No.:  2.11.4
                                                                Date:  January 1990
                                                                Page:  2
                  TABLE 4.1.   SUMMARY OF FILTER ACCEPTANCE CRITERIA
                 40 CFR 50, Appendix J
Explanation
Collection    Greater than 99 percent as
efficiency    measured by the dioctyl
              phthalate (DOP) test, with
              0.3 pm particles at the
              sampler's operating face
              velocity.

Integrity     5 /ig/m3, measured as the
              concentration equivalent
              corresponding to the differ-
              ence between the initial and
              final weights of the filter,
              assuming a 24-hr sample
              volume of 1600 m3.
Alkalinity    Less than 25 microequiva-
              lents/gram of filter.
                                              The apparatus needed to perform
                                              this test is not available for a
                                              typical  analytical  laboratory.  The
                                              operating agency must insure that
                                              the filter manufacturer has
                                              complied with this  guideline.

                                              During a simulated  sampling test, .
                                              all sampling procedures are fol-
                                              lowed EXCEPT the HV PM10 sampler is
                                              not turned on.  The tare weight of
                                              the equilibrated- filter must agree
                                              within ±8 mg of final weight.
                                              Refer to Subsection 3, Field Opera-
                                              tions, for recommended HV PM10
                                              sampling procedures.
                                              A typical'analytical laboratory
                                              capable of conducting this test
                                              measurement of alkalinity  (see"
                                              Reference 8).
                    is

-------
                                                                Section  No.: 2.11.4
                                                                Date:  January 1990
                                                                Page:  3
  aanrpn   J6 QUartZ f11ter'  consi'stency tn labeling these  filters will allow the
  referenc?nPn nf°^ S'V^T  t0/he filter ID number *or documentation and cross-
  "nq thTfmer c^t^fo^  1°™*' *ThiS V" als° e1imi'nate confusion in ?oad-
  ImhnccL K  ?u cassettes  for  subsequent sampling.  If the filter  ID number is to be
  SSaoP  anX   f 0Peratln9 a9ency, gentle pressure must be used to avSid ffuer
  damage, and extreme care  must be taken to avoid  duplication or missed numbers
      Visual  Filter  Inspection
ing  iii                                                n
contaimng numerous defects should be returned to the supplier

     The following are specific defects  to  look for:

     K    PJ5ho]e--A small hole appearing as a distinct and obvious  briqht

                 '
                                                                               of
                                                                               01
                                                          th.t .<,ht be evidence

     '•
     *
4-3  Filter Equilibration
                                                   *   1co
                                                         'fc

-------
                                                                Section No.: 2.11.4
                                                                Date: January 1990
                                                                Page: 4


4.4  Initial Weighing Procedures (Tare Height)

     Enough filters-to last for at least a 3-mo sampling period should be numbered
and weighed at one time.

     Filters must be weighed on an analytical balance with a minimum resolution of
0.1 mg and a precision of 0.5 mg.  Each balance used in the weighing procedures
must be identified by a balance number.  Each balance should be assigned a block of
filter numbers to be used sequentially.  Procedures are as follows:

     1.   Make sure that the balance has been calibrated (at least annually) and
          maintained according to manufacturer's recommendations.  If the balance
          is out of calibration, have it calibrated according to manufacturer's
          directions.

     2.   .Zero the balance according to manufacturer's directions.

     3.   Perform a QC "Standard Weight" check on the analytical balance.   (Proce-
          dures are outlined in the next subsection.).

     4.   If filters must be weighed outside the conditioning chamber, take care to
          avoid interference with ambient hygroscopic particles, and begin the
          weighing procedure within 30 s.  Weigh the filter according to manufac-
          turer's instructions, making sure that a stable reading is obtained.  At
          'routine intervals, check the zero and calibration of the balance as out-
          lined in the next subsection.

          Note;  Be careful when loading and unloading the balance with quartz
          TTTters.  The corners and edges of the filter should not bump the balance
          door because, the filter may break or filter material may be lost.

     5.   Place the tared filter, with the filter ID number facing upwards, in its
          original container or a comparably sized box.  Place a sheet of
          8-1/2 x 11-in. tracing paper between each filter.

     6.   Record the balance number, the filter ID number, and the tare weight on
          the Laboratory Data/Coding Form (see Figure 4.1) or alternative data
          recording medium.  When bound, these forms serve as a laboratory  note-
          book.  Any filter'weight outside of the normal range of 3.7 to 4.7 g
          should be investigated immediately.  Sequentially number each form in the
          upper right-hand corner.

     7.   Perform the tare and gross weight QC checks.as detailed in the following
          subsection and submit all data to the QC supervisor for review.   (Minimum
          QC supervisor activities are also presented in Subsection 4.5.)

4.5  Internal QC

     During the filter weighing process, the following QC checks are recommended.
All QC data, including the actual and measured weights, the date, and the opera-
tor's  initials, should be recorded on an Internal Quality Control Log Sheet  (see
Figure 4.2).

-------
                                                    Section  No.: 2.11.4
                                                    Date:  January  1990
                                                    Page:  5
                         Quality Control
                         Supervisor
Figure 4.1. Example laboratory data/coding form.

-------
Section  No.: 2.11.4
Date: January 1990
Page: 6








N
CO
3-
L.
O)
Q
Balance

G
I
1
/? o
O jrt
» Q)
CO Q.
a co



2:
1
cv
Balance
Operator
en
u
o>
U
S
J?
M
05
2
a
a
1
o
0)
if
"<5
U

o "3
jc £
2 *
, 
O
Is
I1
Observed
Value
• 3
^^
—
To
Q
^
^^^
^^*
^^^
PI\
^^
I
r-
in
^>
0
0
q
S
Q
o<=
W
c^*o
D^
Q

-------
                                                                  Section No.:  2.11.4

                                                                  Date:  January 1990
                                                                  Page:  7



                                                                              -
Control Log Sheet to the laboratorv nr 1™ h««i  V °;  f   °" the Inte™>  Quality
tained on each balance and also °ncl2Sed ?n ?he'loS°^7  'r^5  ^"^  be "ain-
cate any excess drift caused by an inltJent Ll?Snc??on.   ""'  ^^  ""'  1ndi-
and rIcordedTdraVJa?r ""^ dai'ly C6rtify  the  ac«ptability  of  all  filter weights
4'6  P°'st-Samiiling Documentation  and
adhereUP?S tfl]?"  ^ the "•""- the »«P'« custodian shou,d
                                                          ai
            field operator or if a

-------
                                                                Section  No.:  2.11.4
                                                                Date:  January 1990
                                                                Page:  8


     2.    If the exposed filter was  packaged  for shipment,  remove  the  filter  from
          its protective envelope and examine the shipping  envelope.   If sample
          material  has been dislodged from a  filter,  recover  as  much as  possible  by
          brushing  it from the envelope onto  the deposit  on the  filter with a soft
          camel's-hair brush.

     3.    Match  the filter ID  number with  the correct laboratory data/coding  form
          on which  the original balance ID number,  filter ID  number, filter tare
          weight, and other information are inscribed.  The sample custodian  should
          group  filters according to their recorded balance ID numbers.   Initial
          separation of filters by balance ID number will  decrease the probability
          of a balance error that could result from the use of different balances
          for tare  and gross weights.                                   :

     4.    Remove the filter from the protective manila  folder.  Should the filter
          be retained in its filter cassette, loosen the  nuts on the top and  remove
          the filter.  Overtightening of the  nuts may cause the  filter to adhere  to
          the cassette gasket.  Gently remove it by the extreme  corners  to avoid
          damage. .Inspect the filters for any damage that  may have occurred  during
          sampling.  Conduct a secondary'check of a sample's  validity  (as presented
          in Subsection 3.3).   If insects  are embedded  in the sample deposit,
          remove them with Teflon-tipped tweezers and disturb as little  of the
          sample deposit as possible.  If  more than 10  insects are observed,  refer
          the sample to the supervisor for a  decision on  acceptance or rejection  of
        ' the filter for analysis.    ,                                            .

     5.    Place  defect-free filters  in protective envelopes and  forward  them  to the
          laboratory for weighing and analysis.  File the data sheets  for subse-
          quent  mass concentration calculations.

     6.    Place  defective filters, with the type of defect(s) listed,  in separate
          clean  envelopes, label the envelopes, and submit  them to the laboratory
          supervisor for final approval of filter validity.

4.7  Final Weighing Procedure (Gross Weight)

     1.    Place  the defect-free filter(s)  in  a conditioning environment  and  allow
          them to equilibrate according to procedures outlined in  Subsection  4.3.

     2.    Repeat Steps 1 through 6 of the HV  PM10 filter  tare-weighing procedure
          (Subsection 4.4).  Record the indicated gross weight on  the  Laboratory
          Data/Coding "Form.

     3.    Perform the internal QC checks described in Subsection 4.5  to  ensure the
          validity of reweighing.

     4.    If the HV PM10 filter is not to receive additional  analysis, p-1ace  it
          into a protective envelope or folder.  Deliver  weighed filters to  the
          sample custodian for archiving.

-------
                                                                  Section  No.:  2.11.4
                                                                  Date:  January 1990
                                                                  Page:  9


4-8  Calculation of PHIQ  net  Filter  Loading


of PMlSefo'r0tShSae1nlIetret      "  "V  PM1° filter

-------
Section No.:  2.11.4
Date: January 1990
Page: 10













>*
Ul
REPARATION AND ANALYSIS
a.
ce l
s
>-*
1

CM

«r
UJ

*s
H










i
3

k
§4.
kg
.- u
4f

0
4>
U




c
t

••
•
o
k
1

1
I



Requirements
ec
II
filters in protective <
or boxes and number set
1-
! ?, 2~






x
u
c
• a. e


^
i.
3

U
6
e i k

• 0
L.

TJ
C
M

^n
o

C

TJ
C
JC

£
%
-*J 1 U
4! S
i ~~ ! -O
t: i °













j


-

^i ?
•». f .-



._ i _
49 " 1^
U
II X
C



t.
49
%.

U
u
n
2




3"
"o
JC
c
"o.
0
c
k
k
1 .
U
M ^
9 1.
JC •
2 49

L
49


*-
X
u

9
4>
u

o.
n
e

^
M*

*
3
>





JC
U

JC
u

—
3

C ,
o
n
u
49
c
e
"O
t.
0.
o
u
Q.
* JC
2



1
• 1.
'Ilj
JO ~" 49
§.2 13
c 49 e
8 §6
i.
1. T) O
49 -O
— k a
— o c
»- 0 —
• t. o
J= U
49 -O ^.
.Si
k •
e — t'-S
je "a o —
• • 49 .-
Z — m «•-


u


— .
*•

u

•o
jc e
u c
e o
u n
.M
X •


• O
3 -0
« E
— 3
> C



C
o

^

u
«».

t%
£
e


0
3
•
J3
3
o-
e
4>
Q.
ce



i
c
o
u
c
13
O
k
49
C
O
U
c
£
k
J3
—
*3





e
j=

^g
0
u

k
•o
c


0

u
n
0




c
o


J

^ ,
•••
3
O"
UJ













^
e
CM
|
J3
ce
CM
/M
k
0
^
C
u

1


u
k
o

e
o
k

e
o

^
k


H
3
or


A
, n
o
k
a

c

&
^
•






•






1 C
k e
e •  e u> •


0
k
e e
j: Q.
• 0

k 49
« k
^3 ^D e
o — —
. = |
• cr «
e •
0 -C
•^ E o

•^ E C
C C k
0 — 0
U 6 «•-





^^
c
o

u
u

M
e
•o
1
JO
3
I
k
.5
•1
,
O»
|J
K 49



49 k •
12-S
k —
• •>• 0
C k
o -c
49 49 —
49 0 >
.c c g •
a— E k
e k •
k * e -c
e 49 u
49 C 0-
— — • e
— o o
•; o
k — k
« o a
n k
_O 3 >_>
o •« cr




O}
c
.^
-C
as e
e 3
* -o
e
e u
k 0
• k
H- a.
n
e
a
M O
It
13 —
49 k
u e
• a.
H. 3
o
• 49
5 49
•o o
^ c

• «
e
o
• 49
M •
e 49 « .
— con
O f C H
j= £ « o
e 3 -o —
a o > e
-o e —
o a.
c e o E
49 C •
e e •
k — •
-r- <>
e o s
k U k C
e • o
_C .- j; .-
49 • 49 49
U U
• 49 • C
k e k 3
3 3 •«•
* • « —
II •
e k e E
je • -x
2 -2 e "o


B
^

i
* •

^_ *
— c
• 0
•—

u «
^: c
g
X 3
— O
— O
• TJ
3
n 13
— c
1
c
•~ 1
3 13
O U C
C 0 • C
— -o • o

a - c 49
£ C O •
• o — o

1 49 • •».
49 U 4) —
n C c L.
o a. c c
a. « E >
i
k
—
3
C
49
*
_
o>
*• c
* o
ce 49




51 _:
Sr 2
?7-
j= •
O 49 >
49 Q
49?!
J= — k
CSjC
— a u
* H-
• — »
49 CB
J3 « S
O -fi t*'


O»
C
•••
JZ
a
— — «i
* "• k
3
k E -o
e k e
4> O U
— ••• o
— k k
«- • a.
a.
A U
j: -or
49 «
k —
e 3 •
> ^ c
k C k
e u c
« 0 4)
J3 k C
o a.—



a>
c
•—
^
o>
'e k
* 3

« e
« u
o o
k k
o a.















k
•
J3
•
O .
C
o
• ! '
-O




















i!
M
i i
:












-------
                                                           Section No.: 2.11.5
                                                           Date: January. 1990
                                                           Page: 1
2.11.5  CALCULATIONS,  VALIDATIONS, AND REPORTING OF  PM10 DATA

Ho*0,M?aSUr?T"tS °f PI210 mass concentration in the  atmosphere that are used to
determine attainment of the National Ambient Air Quality Standards for participa
 3 '               "*1 1" ""^ °f micr°9rams P«r  standard ?ub?c mete? &/"«
                   e measurements, "standard" means EPA-standard conditions of
                                   •
        air    o
             • 2                  ,                   -r  conons o
           'and pr?ssure'  which are 25 •€ (298 K) and  760 mm Hg  (101 k?a)  ?espec-
            sS^ifffSfsss ::; sal
      Particle size discrimination  by Inertia! separation requires that  soecific  air
 "a                                                        "
and  barometric pressure  (Tav aUd P^rfo? the slmpU JSr,'« j*ual "era3e temperature

-------
                                                                Section No.: 2.11.5
                                                                Date: January 1990
                                                                Page: 2
5.1  Calculations
5.1.1.  Flow Rate Calculations - Because flow control methods (and hence, calibra-
tion procedures) vary among different sampler models, the calculations necessary to
determine the average actual flow rate during a sample run will also differ.  The
following general procedures are recommended for calculating the average ambient
flow rate of the HV PM10 sampler.  In this subsection, it is assumed that the sam-
plers have been calibrated according to procedures outl.ined in Subsection 2.

     Note:  Consistency in units is required.  Adoption of uniform designations of
K for temperature and mm Hg (or kPa) for pressure is recommended in all calcula-
tions.

HFC Sampler - The average actual flow rate for the sample period is calculated by
determining (a) the average of the initial and final manometer readings (APex) [or
the average flow recorder trace], (b) the average ambient temperature  (Tav), and
(c) the average ambient barometric pressure  (Pav) during the sampling  period and
applying these values to the calibration relationship.

     Each sampler's flow measurement system  should be calibrated periodically, and
the calibration should be described by a mathematical expression (e.g., a least-
squares linear regression equation) that indicates the slope and intercept of the
calibration relationship.  Following the procedure in Subsection 2, this expression
is in the'form of:
                        Qa = {[APex(Tav+30)/Pav]1/2 - b} {1/m}
 (Eq.  9)
     where:
            Qa = the sampler's average actual flow rate for the sample period,
          	   m^/min
          APex s average of initial and final sampler manometer readings,  (APexi +
                . APexf)/2, mm  (or  In.) H20
           Tav s average ambient temperature for the sample period, K  (K = °C +
                 273)
           Pav = average barometric pressure for the sample period, mm Hg  (or kPa)
             b s intercept of  the  sampler calibration relationship
             m = slope of the  sampler calibration relationship.-
      [For the flow recorder,

                         Qa  =  {7[(Tav+30)/Pav]l/2  .

     where:
(Eq.  9a)
           I  =  average  flow  recorder  reading  for the  sample period.]

     The  average  actual  flow  rate  is  then corrected  to  EPA-standard  conditions,
calculated as:

-------
                                                              Section  No.: 2.11.5
                                                              Date:  January 1990
                                                              Page:  3



                          Qstd = Qa(Pav/Pstd)(Tstd/Tav)                   (Eq.  la)

     where:

          Qstd  =  average sampler flow rate corrected to EPA-standard volume flow
           _    rate units,  std.  m-Vmin
          P
-------
                                                                Section No.: 2.11.5
                                                                Date: January 1990
                                                                Page: 4
                 Qa = average actual sampler flow rate for the sample period,
                      m^/min  '•
               Pstd = standard barometric pressure,  760 mm Hg (or 101 kPa)
               Tstd - standard temperature, 298 K

5.1.2  Calculation of PM10 Concentrations - Accurate reporting of total PM10 mass
concentration data requires the calculation of the total standard volume of air-
sampled (Equation 21) and the final computation of total PM10 mass concentration
(Equation'22).

     1.  • Calculate the total standard volume of air sampled:


                                 Vstd = (Qstd)(t)                          (Eq. 21)

          where:                '       '       ;

               Vstd = total volume of air sampled in standard volume units, std. m^

               Qstd = average sampler flow rate corrected to EPA-standard condi-
                      tions, std. m3/min
                  t = total .elapsed sampling time, min.

     2.   Calculate total PM10 mass concentration in ^g/std. m3:

                           •  PM10 = (105)(Wg - Wt)/Vstd                    (Eq. 22)

          where:                   .

               PM10 = PM10 mass concentration, /jg/std. m3
                10^ = conversion factor, pg/g
             Wg, Wt = gross and tare weights of the HV PM10 filter,  respectively, g
               Vstd s total sample volume in standard volume units,  std. m3


5.2  Calculation Validation

     Data that are needed to compute the mass concentration of PM10 originate from
two main sources:  field operations and laboratory operations.  These data must be
validated to ensure that all reported PM10 measurements are accurate relative to
the overall scope of the quality assurance program.   When the final  mass concentra-
tion of PM10 in a sample has been computed, the validation procedure not only will
check on these computations, but also will aid in the flagging of questionable mass
concentrations (i.e., extremely high or low values).  Therefore, should a mass
concentration approach the primary or secondary ambient air quality standard, this
validation procedure will provide checks for all preliminary field and laboratory
operations.  The steps of the calculation validation procedure are as follows:

     1.   Gather the following data for each sample:

          •    Total sampling time (min)            	
          •    Average actual volumetric flow rate,  Qa (m3/min)
          •    Tare and gross weights, Wt and Wg, of the HV PM10 filter (g)

-------
                                                                 Section No.:  2.11.5
                                                                 Date:  January 1990
                                                                 Page:  5
           Recalculate. the total mass concentration of PM10 for 7 samples  per 100
           (minimum of 4 per lot) as specified in Subsections  5.1.1  or 5.1.2.   These
           «;Sr?!nr fre;ju?nc1es may be ^justed subsequently,  based on accumulated
           experience and level  of data quality.  Decrease  the  frequency if  experi-
           ence indicates that data are of good quality,  or increase it if data are
           OT  marginal  or poor quality.  It is more important to  be  sure that  the
           validation check  is representative  of the various conditions that may
           influence data quality than -to adhere to a fixed frequency.          '

           Compare  each validated PM10  concentration with the originally reported

           datfof  rorrprJ'any ^"V^*  are f°und'  initia1 the^  and  Indicate the
           tinna? «?rM» T ' i    *  ^  Per(?enta9e of errors  is found, check  addi-
           tional calculated values.  If  consistent  errors  are  found  check all val-
           ues  in the block of data  and investigate  and correct the caSsI
     4'   f?3!!31^.1?431 TaSS c°ncentration values; note those that appear exces-
          sively high or low and investigate.  Repeat Steps 2 and 3 for these sam-

     5'   Mnh1! m?SS C0(?centrat|on computations appear correct and questionably
          high or low values still exist, review all raw data (i.e.  sample time
          HSrH'^^-J^Tr1'0 fl°W rate''and its subsequent  orrectP on to '
          standard conditions) for completeness and correctness.

5.3 • Data Reporting  .  .
on     ,        °f the standards for particulate matter in the ambient air is  based
?ation of PMW JSl3 c°"centrati°n °fuPM1°-   Inf°nnation on report  ng  and  interlrl-
5S C?R ?J,  Appendix K    reSf>eCt t0 the attainme"t of these standard!  is covered  in

-------
                                                                 Section No.: 2.11.5
                                                                 Date: January 1990
                                                                 Page: 6
                 TABLE 5.1.  FORMULAS ASSOCIATED WITH PM10 MONITORING
      Calculation
               Formula
 Equation
 Conversion of flow rate
 from actual to standard
 volume units

 Conversion of average flow
 rate from actual  to stand-
 ard volume units

 Conversion of flow rate
 from standard to  actual
 volume units

 Uncorrected air volume
 measured  by standard
 volume meter

 Correction of air volume
 measured  by std.  vol.  meter
 to  ambient baro.  pressure

 Actual  volumetric flow
 rate measured by  standard
 volume meter

 Actual  volumetric flow
 rate measured by  orifice
 transfer  standard

 Transformed  exit  orifice
 pressure  for MFC  sampler
 calibration  relationship

 Transformed  flow  recorder
 reading for  MFC sampler
 calibration  relationship

 Regression model  (y=mx+b)
 for calibration of MFC
 sampler

Regression model  (y=mx+b)
for calibration of MFC
sampler using flow recorder
    Qstd  = Qa(Pa/Pstd)(Tstd/Ta)
   Qstd = Qa(Pav/Pstd)(Tstd/Tav)
   Qa = Qst'd(Pstd/Pa)(Ta/Tstd)
 (Eq.  la)
 (Eq. 2)
   AVol. '= Final Volume - Initial Volume   (Eq. 3)
   Va = AVol.(Pa - AHg)/Pa
   Qa = Va/ATime
(Eq. 4)
(Eq. 5)
Qa(orifice) = {[AH20(Ta/Pa)]l/2 - b} {1/m} (Eq.  6)
   APext = [APex(Ta+30)/Pa]l/2
   It = I[(Ta+30)/Pa]l/2
   APext  =  m[Qa(orifice)]  + b
   It  =  m[Qa(orifice)]  +  b
(Eq. 7)
(Eq. 7a)
(Eq.  8)
(Eq.  8a)
                                                                      (continued)

-------
                                                                  Section No.:  2.11.5
                                                                  Date:  January 1990
                                                                  Page:  7
      Calculation
TABLE 5.1.  FORMULAS ASSOCIATED WITH PM10 MONITORING (Cont'd)

                                Formula                      Equation
 Calibration relationship.
 for MFC sampler

 Calibration relationship .
 for MFC sampler using
 flow recorder

 Set-point flow rate for
 MFC sampler

 Set-point manometer
 reading for MFC.sampler

 Set-point reading for
 MFC sampler using flow
 recorder

 Conversion of  manometer
 reading in inches  of
 H20 to mm Hg

 Absolute stagnation
 pressure

 Stagnation press,  ratio

 Avg. stagnation  press,  ratio

 Regression model  (y=mx+b)
 for calibration  of VFC
 sampler

 Calibration  relationship
 for VFC  sampler

 Audit or-QC  flow check
 of sampler calibration

 Audit or QC  flow check
 of sampler operational
 flow rate

 Regression model  (y=mx+b)
 for provisional calibration
of VFC  sampler  for audit
                     Qa = {[APex(Tav+30)/Pav)]l/2 . b} {1/m} (Eq. 9)


                     Qa = {T[(Tav+30)/Pav]l/2 - b} {1/m}     (Eq. 9a)
                     SFR  =  (1.13)(Ps/Pa)(Ta/Ts)
                     SSP  =  [Pa/(Ta  + 30)] [m(SFR)' + b]2


                     SSP  =  [m(SFR)  + b][Pa/(Ta+30)]l/2
                                      (Eq,
                                      (Eq.
                                      (Eq.
      10)


      11)


      lla)
                    mm
Hg = 25.4(in. H20/13.6)
                                                             (Eq.  12)-
                    Pl = Pa - APstg. PI = Pa-- APstg

                    Stagnation press, ratio = Pi/Pa

                    Avg. stag, press, ratio = "PT/Pav


                    Pi/Pa = m[Qa(orifice)]/[Ta]l/2 +
                                     (Eq.

                                     (Eq,

                                     (Eq.
     13)

     14)

     14a)
                                     (Eq. 15)
                Qa(sampler)  = {[(Pl/Pav) - b][Tav]l/2}{i/m} (Eq

                ^ diff.  = [Qa(sampler)  - Qa(auditn  r   i
                          I\Qa(audit)J  [100J
                <  difference = [Qa(audit)  - 1.13
                                       1.13
                           [100]
(Eq.
16)


17)


18)
                    (Pl/'Pa).  =  m'[Qa(audit)]  + b1
                                     (Eq.  19)
                                                                      (continued)

-------
                                                                Section No.: 2.11.5
                                                                Date: January  1990
                                                                Page: 8


            TABLE 5.1.  FORMULAS ASSOCIATED WITH PM10 MONITORING  (Cont'd)

     Calculation                            Formula                      Equation

Provisional calibration of
VFC sampler for audit            Qa(audit) =  [(Pi/Pa) * b']/m'           (Eq. 20)


Total air volume sampled         Vstd = (Qstd)(t)           '             (Eq. 21)

PM10 mass concentration          PM10 = (106)(Wg - Wt)/Vstd              (Eq.- 22)

Corrected sampler flow rate      Qa(corr. sampler)
under normal operating con-
ditions during audits and          Tna/eam«i«J [100 - % difference!     (Eq. 23)
QC flow checks                   s [Qa(sampler] [-ygg•j


                                       SYMBOLS


           b   Intercept of linear regression calibration -relationship

          b'   Intercept of linear regression for provisional  calibration for audit
               of VFC sampler

        AH£0   Pressure drop across a transfer standard orifice,  mm (or in.)  of
               water column

         AHg   Differential pressure at inlet to standard volume  meter,  nan Hg (or
               kPa)

           I   Flow recorder chart reading,  arbitrary units on square-root-function
               scale

           I   Average flow recorder chart reading over the sample period,  arbi-
               trary units on square-root-function scale

          It   Transformed flow recorder reading,  for calibration  relationship

           m   Slope of linear regression calibration relationship

          m1   Slope of linear regression for provisional  calibration for audit of
            .  VFC sampler

          Pa   Current ambient barometric pressure,  mm Hg  -(or  kPa)

         Pav   Average ambient barometric pressure for the sample  period,  mm  Hg (or
               kPa)

        APex   Pressure in exit orifice plenum of sampler,  measured with  respect to
               atmospheric pressure,  mm (or  in,,)  water column

        APex   Average of initial  and final  APex  readings,  mm  (or  in.)  H20

                                                                       (continued)

-------
                                                                Section No.: 2.11.5
                                                                Date: January  1990
                                                                Page: 9
            TABLE 5.1.  FORMULAS ASSOCIATED WITH PM10 MONITORING (Cont'd)
                                  SYMBOLS (cont'd)
       APext

      "  PM10
        Pstd
      APstg

      APstg
          PI
          Pi
      Pi/Pa
       ., Q*
         QS

  Qa(audit)

Qa(orifice)

Qa(sampler)

       Qstd
       Qstd

        SFR

        SSP

         t
         Ta
       Tav
  Transformed  exit  orifice  plenum pressure,  for  calibration  relation-
  snip, mm  (or in.) water column
  PM10 mass concentration,  /Kj/std. n>3                     '     '
  EPA-standard  atmospheric  pressure, 760 mm  Hg (or  101  kPa)
  Relative stagnation pressure, measured with respect to  atmospheric
  pressure, mm  Hg
  Average of initial and final APstg readings, mm Hg
  Absolute stagnation pressure, mm Hg
 Average absolute stagnation pressure for the 'sample period, 'mm Hg
 Ratio of absolute stagnation pressure to current barometric pressure
 Average ratio of absolute stagnation pressure to barometric pressure
 for the sampler period
 Sampler flow rate measured in actual  volumetric units, m3/min
 Average sampler f Tow rate for the  sample period measured in actual
 volumetric  units,  nH/min
 Sampler flow rate in  actual  volumetric  units determined  by  a flow
 rate  audit, . nH/min
 Flow  rate measured by  an  orifice transfer standard in  actual
 volumetric units,  m3/min
 Flow  rate  in  actual volumetric units  indicated  by  sampler's calibra-
 tion  relationship  during  flow rate  audit  or QC  flow check,  m3/min
.Flow  rate measured in  EPA-standard  volumetric units, std. m3/min

                           f°r the Sanp1e  peHod  in  standard
Set-point Mow rate in actual volumetric units for MFC sampler
m-Vnnn                            .                   .     r    '
MFr™n        * mf°m!ter reading [or flow recorder reading] for
MFC sampler needed to obtain SFR, mm Hg [or I]
Total, elapsed sampling time, min
Current ambient temperature, K (K = °C + 273)
Average ambient temperature during the sample period, K
                                                                      (continued)

-------
                                                          Section No.:  2.11.5
                                                          Date: January 1990
                                                          Page: 10

      TABLE 5.1.  FORMULAS ASSOCIATED WITH PM10 MONITORING (Cont'd)

                            SYMBOLS (cont'd)

 ATime   Elapsed time during which a flow rate is measured by a standard
         volume meter, min                                    -       •
  Tstd   EPA-standard temperature, 298 K
    Va   Actual air volume measured by standard volume meter at ambient tem-
         perature and barometric pressure, m^
  AVol   Uncorrected air volume measured by standard volume meter, m^
  Vstd   Total sample volume measured in standard volume units, m^
Wg, Wt   Gross and tare weights of the HV PM10 filter, respectively, g

-------
                                                         Section No.: 2.11.6
                                                         Date: January 1990
                                                         Page: 1
 2.11.6 MAINTENANCE
      Maintenance is defined as a program of positive  actions aimed toward prevent-
  ing failure of monitoring and analytical systems.  The overall objective of a rou-
  tine preventive maintenance program is to increase measurement system reliability
  and to provide for more complete data acquisition.                 -

      This section outlines general  maintenance procedures for HV PM10 samplers
         r?IIP  Jfr ^ °rjatl'on °n a  Particular sampler or on laboratory equipment
                     h  mnufaturer's instruction manual for the individual
                              °f thiS Section' su™-izes maintenance activities
     Recor^ should  be maintained for the maintenance schedule of each HV PM10
     nl't  ^ ?hou1d ref1e^ the history of maintenance, including all replace-
     parts suppliers, costs, expenditures, and an inventory of on-hand spare
 equipment for each sampler.  Check sheets should be used  to record preventive
 and/or corrective maintenance activities and the subsequent sampler calibration
 6 . 1  Maintenance Procedures
     The HV PM10 sampler is comprised of two basic components:  the inlet and  the
 flow control system.  Because of the differences between sampler models  ft will be
 necessary to refer to the manufacturer's instruction manual for specific  steo-bv-
 step maintenance guidelines and necessary supplies.           speuiTic, step oy- .

 6.2  Recommended Maintenance Schedules
          the following maintenance frequencies.

       !mpact1on Inlet  -.The impaction  inlet  should be dismantled and


^^m%m^s^<"^~-

-------
                                                                Section No.: 2.11.6-
                                                                Date: January 1990
                                                                Page: 2


6.2.3  HFC Base - The MFC base is equipped with the following items:

     1.   Connecting tubing and power lines, which must be checked for crimps,
          cracks, or obstructions on sample recovery days.  Fittings should be
          inspected periodically for cross-threading and tightness.

     2.   A filter screen, which should be inspected on sample recovery days for
          any impacted deposits.

     3.   Filter cassette gasRets,  which need to be inspected each time a cassette
          is loaded.  A worn cassette gasket -is characterized on exposed filters by
          a gradual blending of the'boundary between the collected particulates  and
          the filter border.

     4.   Motor and housing gaskets, which should be .checked at 3-mo intervals  and
          replaced as necessary.

     5.   Blower motor brushes, which should be replaced before they become worn to
          the point that damage may occur.  Although motor brushes usually require
          replacement after 600 to  1,000 h of operation, the optimum replacement
          interval must be determined by experience.  A pumice stone can be used
          against the motor's, contacts to ensure high conductivity.  Change the
          brushes according to manufacturer's instructions,  and perform the opera-
          tor's field-calibration check as presented in Subsection 3.5.  If the
          sampler's indicated flow  rate exceeds the manufacturer-specified design-
          flow-rate range, adjust the sampler before the next run day.

          To achieve the best performance, new brushes should be properly seated on
          the motor's commutator before full  voltage is applied to them.  After  the
          brushes have been changed, operate the sampler at  50 to 75 percent of
          normal  line voltage for approximately 30 min.  The motor should return to
          full  performance after an additional  30 to 45 min  at normal  line voltage.

          Caution;  The motors that are used for HV PM10 samplers are higher-cur-
          rent  versions of the motors that have been used for HV total  suspended
          particulate samplers.  The brushes  for the two types of motor are differ-
          ent.   Make sure.that the  correct replacement brushes are used for the
          maintenance of HV PM10 samplers.

     6.   If a  motor needs to be replaced, be sure to use the higher-current ver-
          sions that are needed for HV PM10 sampling.   When  lower-current motors
          are installed in- HV PM10  samplers,  the flow rate has been found to vary
          with  changes in the line  voltage.

     7.   A flow controller,  which  should  be  replaced if the flow recorder indi-
          cates no flow,  low flow,  excessive  flow,  or erratic flow.  'Minor adjust-
          ments can be made to alter sampling flow rates;  however,  the  controller
          generally cannot be repaired in  the field.

     8.   A flow recorder,  which  requires  very  'little  maintenance,  but  does  deteri-
          orate with age.   Difficulty in  zeroing the recorder and/or sign-ificant

-------
                                                                Section No.: 2.11.6
                                                                Date: January 1990
                                                                Page: 3
          differences (i.e.,  greater than 0.3 m3/nrin)  in average flow rates ob-
          tained from consecutive sampling periods usually indicate a faulty re-
          corder.  The recorder pens should be replaced  every 30 recording days
          In dry climates,  a  more frequent replacement  schedule may be required.

6.2.4  VFC Base - The VFC base is equipped with  the following items:

          Power lines,  which  must be checked  for crimps  or cracks  on  sample recov-
          ery days   Fittings should be  inspected periodically  for cross-threading
 1.

     and tightness.
     2.    A  filter  screen  and  the throat of the choked-flow venturi,.which should
          be inspected  on  sample recovery days for any impacted deposits.

     3.    Filter cassette  gaskets, which should be checked each time a filter is
          Innr^ f?\i  H°rn CfSuUu 9asket is characterized on exposed filters by
          a  gradual blending of the boundary between the collected particulates and
     the filter border.

4.


5.
         ?nrSni!?t?ha?rHSheS| Whl'Ch Sh°Uld be reP]aced before they become worn to
         the point that damage may occur.  Although motor brushes usually requir*

              "        ^0       " °f °Peration' ^ °P«™« replLS
                  m st bede                          '
          nt tKo   J  aetenmned by experience.   A pumice stone can be used
         against the motor's, contacts to ensure high conductivity.   Change the
         ?Jr'sef,>l5°rJr8 to.»anu^turer's instructions,  and perform the opera-
         tor s f eld-calibration check as presented  in Subsection  3.5.   If the
         sampler s indicated flow rate exceeds  the manufacturer-specified  desiqn-
         f low-rate range,  recalibrate the sampler before  the  next  run  day.     9

         To achieve the best performance,  new brushes  should  be  properly seated on
         the motor's  commutator  before full  voltage  is applied tb them   After the
         brushes have been changed,  operate  the sampler at  50  to 75  wrcent  Sf
         normal  line  voltage for approximately  30 min.  The motor sho^S retu  n to
         full  performance  after  an  additional. 30  to  45 min  at  normal line  voltage?

                   The  motors  that  are  used  for HV PM10 samplers are hiaher-
                                         replacement brush
        mointenance or HV PM10 samplers.

   6.   If ,
                        5e/ePIaced.  »e sure to use the higher-current
                                                                           ver-

-------
                                                                Section No.:  2.11.6
                                                                Date: January 1990
                                                                Page: 4
6.3  Refurbishment of HV PM10 Samplers
     If operated in the field for extended periods, HV PM10 samplers may require
major repairs or complete refurbishment.  If so,  refer to the manufacturer's in-
strument manual before work is undertaken.  A sampler that has undergone major
repairs or refurbishment must be leak-checked and calibrated prior to sample col-
lection.

-------
Section No.: 2.11.6
Date: January 1990
Page: 5











CO
UJ
»«(
r"
^
HH
I—
UTINE MAINTENANCE AC
o
CE
CO
LU
m
t-







•
ti
, 49
C
9
' V 49
-I
> 3
. O"*>
a o
1. C
•— c
c
o
49
O






tl
49
"i
a
o
c
a.
o
u



Frequency and/or
method
i
CL
"3
er
.
"?
a d>
:!
O tl
— a
a L. •
a o
«
-.0
c
a 4>
a c
w I



a
^
a

3 a
o a
49 E
t. m
m -o .
o.
L.
tl O
3
O •
— 49
-O *•
0 0
a
o a
Z -o


49 T)
a c
Dismantle and clean
manufacturer-spec i f
interva Is.



1 49
i • •
.49 .f i
qu i pmen
t. c
! r
LU to CO





£
a
ti
ti
o
c
tl
u
a
CE






a
E
•o
0
o
o

II
o
L,
O
u
o
ti
a.
u
0
L
O
•*•
-X
0
a
-c



n
a
c
L.
a
*
0
a.







-

c
a
"
c
a
a
^
u


No obvious depositi
with wire brush.


1 .
I
Visual ly check on si
recovery days.


c
e
a
L. 49
O •
• 0
U
t- JC
a 4>
— c
u. •






X
L
tl
tl
a
o
c
tl
u
a
a
*

c
o
•~
**

a
c
o
O 49
o a
c -o
•T a
z -5

49 L
o. a
- a
• -c
— 49
> e
K
C 49
JC
?a
m
49 —
< a





Gasket






j-
ti
•
a
o
a
c
J
a
o
a
a
et






a
49
L.
j
a
49
CO



O C
49 0
S 49
IS "
« J.
L. a
a o
a -c
o
as
a -

n
ti
o
0.
•^
repair
L.
O
O
a
*

1
CD
3
O
L
_c
49
49 •
• C
1. 3
o a
:r
J3
• 49
49 3
in o



Check when flow-rate
changes are evident.

L.
a
o
c
0
*
0
u.

_5
Ji
tl
o
a.
v.
repair '
t.
. o
o
a
CL
£

•

— c

a
Recordar stays zaro
chart advances;' pen

_
QUO
C 0 —
u -
c o c
Inspect when expert e
•difficulty in zero in
when large changes i
•e
, u
*—
c
•o
a
c
•o
i.
o
o
e






















rates occur.












X
a
ti
a
u
c
tl
u
Q.
cc


1
tl
t- tl
o o
No crimps, cracks,
obstructions; no cr
threading.

i
a
"a.
Visual ly inspect on
recovery days.
tl :
CB
C
"jj
=
CO
c
IZ


-------

-------
                                                                  Section No.: 2.11.7
                                                                  Date: January  1990
                                                                  Page: 1
 2.11.7  AUDITING  PROCEDURES
      The operating agency must perform QA audits and process evaluations to deter-
 mine the accuracy of the PM10 monitoring system and, hence, the data it produces
 The primary goal of an auditing program is to identify system errors that may re-
 sult in suspect or invalid data.  The efficiency of the monitoring system (i.e.
 labor input vs. valid data output) is contingent upon effective QA activities  '
 This true assessment of the accuracy and efficiency of the PM10 measurement system
 can only be achieved by conducting an audit under the following guidelines:     .

      •    Without special preparation or adjustment of the system to be audited.

           By an individual  with a thorough knowledge of the instrument or process
           being evaluated,  but not by the routine operator.

           With accurate,  calibrated,  NIST-traceable transfer standards that are
           completely independent of those used for routine calibration and  OC flow
           checks.

           With complete documentation of  audit information for  submission to the
           operating  agency.   The audit information  includes,  but  is  not  limited to
           types of  instruments and audit  transfer  standards,  instrument  model  and  '
           serial  numbers, transfer-standard  traceability,  calibration  information,
           and  collected  audit  data.            •
           n     ™n        d?scrib*d  in  this  subsection  produce  two  quantitative
          tf  3     °  !fP1er s  Performance:   The  audit flow  rate percentage differ-
                31?"    °W  ra!6  Percenta9e  difference.  The audit flow  rate percent-
        fere"ce. ^ermines  the  accuracy  of the  sampler's  indicated flow rate by
 r     nir?  \  W1tj  *  flow  rate  from the  audit transfer standard.  The  design flow
 the  iS?ete2eSa?an f(nere^e deHtenTn'nes "?0w c1osely the sampler's  flow  rate.matches
 tne  inlet  design flow  rate under normal operational conditions.
                    obser^r should be Pres^t for the audit, preferably the rou-
 nteaitv o  th   ,  -*5*"^"9^'11??6 nt' uThis practice -not on^ Contributes to the
 integrity of the audit, but also allows the operator to -of fer anv exDlanations an^
 informat^n that will help the auditor to determine the po sible^auses of 5? creo-
 ancies between audit-standard values and the sampling equipment values.    Q1screp
                                  i  different mod^s of samplers because of dif-
                          rates, flow-controlling devices,  options utilized fi  P
continuous-flow recorder) , -and the configuration of the samplers   The audit orace
dures provided in this section are specific to high-vofuJI fflj )  mil samolers that'

^K^1^^^
     e^

-------
                                                                 Section  No.: 2.11.7
                                                                 Date:  January  1990
                                                                 Page:  2


7.1  Flow-Rate  Performance  Audit  Procedure  for  Mass-Flow-Controlled
     (MFC) HV PM10  Samplers

     For this MFC procedure, the  following  conditions  are  assumed:

     •    The MFC sampler utilizes an electronic mass-flow controller  for flow-rate
          control.

     •    The sampler's flow rate.is measured by a water or oil manometer connected
          to the exit orifice plenum pressure port [or, if necessary,  by a continu-
          ous flow  recorder connected to the exit orifice  port and equipped with
          square-root-scale chart paper].

     •    The sampler inlet is designed to  operate at  a flow rate of 1.13 m^/min at
          actual conditions; the  acceptable flow-rate  fluctuation range  is ±10 per-
          cent  of this value (i.e., 1.02 to 1.24 m3/min).

     •    The calibrated, NIST-traceable audit  transfer standard is an orifice
          device with an associated water or oil-manometer.

     •    The audit orifice transfer standard'.s calibration relationship is ex-
          pressed in actual volumetric flow-rate units (Qa) as described in Subsec-
          tion 2.2.

     Note:  Do not attempt to audit HV PM10 samplers under windy conditions.
Short-term wind velocity fluctuations will  produce variable pressure readings by
the audit orifice transfer standard's manometer.  The audit will be less precise
because of the pressure variations.

     The auditor should adhere to the following procedures during an audit of the
MFC sampler:

     1.   Transport the following equipment to the monitoring site:

          •    Audit orifice transfer standard with calibration relationship in
               actual volumetric flow-rate  (i.e.,  Qa) units and traceable to NIST
                (see Subsection 2.2).   This orifice transfer standard should not be
               the same one that is used for routine calibrations and QC flow
               checks.

          •    An associated water or oil  manometer,  with a 0- to 400-mm (0- to
               16-in.)  .range and minimum scale divisions  of 2 mm (0.1  in.).

          •    A thermometer,  capable of accurately measuring temperature over an
               appropriate range to the nearest ±1  °C and referenced to an'NIST or
               ASTM thermometer within ±2  °C at least annually.

-------
                                                        Section No.: 2.11.7
                                                        Date: January 1990
                                                        Page: 3
       A portable aneroid barometer  (e.g., a climber's or engineer's altim-
       eter)   capable of accurately  measuring ambient barometric pressure
       over the range of approximately 500 to 800 mm Hg (66 to 106 kPa) to
       the nearest- mm Hg and  referenced within ±5 mm Hg to a barometer of
       known  accuracy at least annually.

       MFC Sampler Audit Data Sheet  such as shown in Figure 7.1 (blank
       forms  appear in  Subsection 12).                 y        ^,cn*

       Clean  filter [and clean recorder chart,  if a flow recorder is used
       to  quantitatively measure the flow rate].
 SfhTr nl^o! V?e-at°r isr«Ponsifale for providing the manometer (or
 other device) that is normally used for measuring the sampler's  flow
 rate  the sampler calibration relationship that is currently in  effect
 for determining the flow rate for sample periods,  and any other  informa-
 tion or equipment that is normally used tb determine ^sampler's  in™-
 cdie                                                    •
  [If a continuous flow recorder is bei'ng used  quantitatively  in  lieu of  a
 s/rsis: *; rare sa;p1r flow rate  record  the  *"*  *°«  ° J  s4ie
 S/N  date, and the auditor's initials  on the  blank  side  of a clean re-
 recordVr c'hart °P?? th* £?"* "T °f  the $ampler and  ^'a^ the clean
 2S?£ rharJ £*    +t    J sampler was calibrated  by  using square-root-
-rh*rt ±H **?£*'  the audlt must be conducted  with the  same type of
 chart paper.   Observe the recorder zero setting.  If necessary  instruct
 the operator to adjust the pen  to indicate true zero.]   55dry' ,instruct
 3'   ,!:!trUSM!!! °P!l?^i:.t0,inSta.11  a.?lean.fi1ter in the HV PM10.  DO NOT
                                                      on the
   ren      at-i;                                  samPer
 screen.   An  audit  filter  should never be used for subsequent semoli

 ?heea?n?e?ais1Cr  ?Ldar9?h  ^^ 1° ^ Can 5? C°11ected on^he ffuSr
 me iniet is raised.  The sampler mass wil  be biased as a result of
 using  a  filter  for both an audit and subsequent sampling.
 Chec^tha^th^ilicJT1" transfe!; standard's faceplate on the sampler.
 Check  that the gaskets are in good condition and have not deteriorated.
                                     nuts evenly on alternate corners  to
                hgn and uniformly seat the gaskets.  The nuts should be  hand-
               only, too much compression can damage the sealing gaskets.

.5.    Install the audit orifice transfer standard with no 'resistance Plate   or
             J va ve of a vanah-ia_^Qcip + ,«^« ----x----  TTP	_  "     '  ul
onnK    ,                                                       e  o
^r* i?ValvVf a variab'le-resistance orifice wide  open.   Forrsist-
ance plate orifices, make sure the orifice gasket  is  present  and the
       riflr       Standard is not  crosf-threadedPon  the
                                Standard'S P-sure port

-------
                                                                          Section No.:  2.11.7
                                                                          Date:  January  1990
                                                                          Page:  4
                            MFC SAMPLER AUDIT DATA SHEET
Station Location


Sampler
                MlLfoe.D
                                  Date
                                                      J/J//39   Time
. mm Hg. Ta
                                   °C
                                        "2-7
                                                  K, Unusual Conditions
                  IO3<2~7
                                             Orifice Calibration Date   /2. //5/ o%
Audit Orifice S/N

Orifice Calibration Relationship:  m =   /• 2.*V3O   b »  ~O- OQ &2- r ~  0.9*777


Sampler Calibration Relationship:  m =   O- 9/37    b =   O-2.g2.*7  r ,  Q. 9977


Orifice Pressure Drop UH2O)   5-*7   in. H2O     Qa (audit)3 	'•  /  '   '	m3/min


                                     With Orifice Installed         Without Orifice Installed

                                        H.O
Sampler Pressure Drop


Qa (sampler)15
                          1. 1
in. H2O


m3/min
                                                                    I-J5J
in. H2O


m3/min
Audit Flow Rate Percentage Difference0


Qa (corrected sampler)01  __ _ I.I//
                                     • _  -7  i_;
                                          £-' '
Design Flow Rate Percentage Difference6
                                                   . m3/min


                                                   	%
aFor calculation of audit orifice standard flow rates:

 Qa (audit) = {U(H2O)(Ta/Pa)r/z - b} {1/m|

''For calculation of sampler flow rates:

 Qa (sampler)  -  {(UPex/PaMTa + 30)I'/2 - b> Ji/mj
"Audit % Difference
                         (sampler) - Qa (orifice) 1 r
                             Qa (orifice)



^Qa (corrected sampler) * Qa (sampler) -
                                                      orifice installed.
                                  11 iuu |  where Qa (sampler) is measured with the




                        100 - Audit % Difference
                                              100
                                                            where Qa (sampler) is measured

                                                            without the orifice installed
-^     „                      fOa (corrected sampler) - 1.13 1 r  ~\
eDesign Flow Rate % Difference =	10°
Auditor
                                                  Observer
                       Figure 7.1.  Example MFC sampler audit data sheet.

-------
                                                             Section No.:  2.11.7
                                                             Date:  January 1990
                                                             Page:  5


  6.    Leak  test  the audit  system (refer to Subsection  2.3.2,  Step  5).   Identify
       and correct  any  leaks  before  continuing.

  7.    Inspect  the  audit  orifice  manometer  connecting tubing  for  crimps  or
       cracks.  Open the  manometer valves fully  and  blow  gently through  the
       tubing,  watching for the free  flow of  the fluid'.   Adjust the  manometer
       rnn   n?  It     -!?*that  the  zero  11ne  is at  the bottom of the meniscuses.
       Connect  the  audit  manometer to  the pressure port on the orifice.  Make
       sure  the unconnected side  of the manometer is open to the  atmosphere.
       Make  sure that the tubing  fits  snugly on  the'pressure port and on the
       manometer.


 8"    to™ min)^  5amp1er  and a11°W U to warm up to °Pei"ating temperature (3


      f71^folhf/al!pl!r*1nlet may be Partially lowered over the audit orifice  "
      transfer standard  to act as a draft shield (if a shield is not otherwise
      provided).,  use a-block to provide at least 2 in. of clearance at thT
      bottom for air flow and for the manometer tubing.
       h6 3nd reC°rd the f°11owin9 Parameters on the MFC Sampler Audit Data
           Sampler location,  date,  time.       •       '
           Sampler model  and-S/N,  and calibration relationship
           Ambient temperature (Ta),  K (K •= °C +'273)
           Ambient barometric pressure (Pa),  mm Hg or'kPa.
           Unusual weather conditions.
           Audit  orifice  transfer  standard  S/N and calibration  information

 10.   When  the  sampler  has warmed  up  to operating temperature,  observe the
      pexor thP rJetrat°r "t0^ead the Samp1er exit orifl'ce "anometer reading,
     data sheet!!  C°ntinuOUS f1ow Border response, I], and record it on the


•12.  Turn off the sampler and remove the audit orifice transfer standard  h.it
     do not remove the filter.  Turn the sampler on again and repeat Step H
     for the normal operating flow rate.                       repeal btep u

13.  Gather together all audit data,  including the audit orifice transfer
     standard's calibration information,  the MFC sampler's ca ibrat?Sn data
     [and the recorder chart that graphically displays  the sampler response].

14.  Verify that the correct readings have been inscribed on the data sheet.


                                                                standard,  ,s

-------
                                                           Section  No.:  2.11.7
                                                           Date:  January 1990
                                                           Page:  6
16.
17.
18.
19.
                 Qa(audit)  = {[AH20(Ta/Pa)]l/2 - b}  {1/m}
                                                                  (Eq.  6)
     where:
     Qa(audit) = actual volumetric flow rate as indicated by the audit
                 orifice transfer standard;  nwmin
          AH20 s pressure drop across the orifice, mm (or in.) H20
            Ta * ambient temperature, K (K = °C + 273)
            Pa = ambient barometric pressure,  mm Hg (or kPa)
             b - intercept of the audit-orifice transfer standard's cali-
                 bration relationship.
             m - slope of the audit orifice, transfer standard's calibra-
                 tion relationship

Instruct the operator to calculate the sampler's indicated flow rate,
Qa(sampler) with and without the orifice installed, as it is normally
done, using the sampler's calibration relationship (Equation 9 [or 9a],
Subsection 2.3.3) and record both Qa(sampler)  values on the data sheet.

Calculate the percentage difference between the sampler's indicated flow
rate, Qa(sampler) with the orifice installed,  and the corresponding audit
flow rate, Qa(audit), determined from the audit orifice transfer standard
     as:
     tadit fl« rate Vdiffere.ee v['"(""''&LilJ)("""t)]
                                                          100
                                                                 (Eq.  17)
                                                                     If
Record the audit flow rate percentage difference on the data sheet.
the audit flow rate percentage difference is less than or equal  to
±7 percent, the sampler calibration is acceptable.  Differences  exceeding
±7 percent require sampler recalibration.  Differences exceeding ±10
percent may result in invalidation of all data subsequent to the last
calibration or valid flow check.  Before invalidating any data,  double-
check the sampler's calibration, the audit orifice transfer standard's
certification, and all calculations.

Calculate the corrected sampler flow rate, Qa(corrected sampler),  using
Equation 23:
     Qa(corr. sampler) = [Qa(sampler)]
                                   100 - audit
                                             100
                                                 difference]
                                                                 (Eq. 23)
     where Qa(sampler) is for the measurement without the audit orifice trans-
     fer standard installed.  Be sure to carry over the sign of the audit %
     difference from Step 17.

     Calculate the design flow rate percentage difference between the cor-
     rected sampler flow rate, Qa(corr. sampler), and the inlet design flow
     rate of 1.13 m^/min as:

-------
                                                               Section No.:  2.11.7
                                                               Dat<••;  January 1990
                                                               Page:  7
          Design flow rate _ fQa(corr.  sampler) - 1.131 r,nrti
            % difference   ' [   - 1.13    ' - ^ [lOOj              (Eq. 18)
2°'  r^rl^6  JeS1'9J- H°W rate. Pontage difference.   If the design flow
     niir KVK  J?e dlfference i? less than or equal to ±7 percent, the sam-
     pler calibration  is acceptable.   Differences  exceeding ±7 percent should
     be  investigated,  and deviations  exceeding  ,10 percent (or the acceptab e
     ?™]?-/]°W~ra*e  ra"3e sP<*lfled fay the inlet manufacturer) may result in
     valfd ??n10rh °V VP °^tai?^ sub^quent  to the last calibration or
     or  ft! £™S?i'«, Bejore(invalld?^ng any  data, double-check the audit
     orifice transfer  standard's certification  and all calculations.

     Note:  Deviations from the inlet design  flow  rate may be caused in  cart
     by  deviations in the ambient temperature and  pressure from the seasonal

             °110n    CCUlatethe  °t        P-
                              C?nCUla
-------
                                                                                 Section  No.:  2.11.7
                                                                                 Date:  January 1990
                                                                                 Page:  8
        Install filter and orifice
      on sampler and leak check
            (Steps  1-8)
       Measure audit flow rate,
             Qa (audit)
           (Steps 9.12)
     Sampler operator measures
      stagnation pressure ratio
      and determines indicated
       flow rate, Qa (sampler)
      from sampler caJbration
         (Steps 10.11.13)
                                             P1/Pa not on
                                             lookup table
         Calculate tentative
          % difference  for
         sampler calibration
         (Steps 14.15.16)
                                  %diff. not within ±6%
 Measure stagnation pressure and
Qa (orifice) at 2 or more additional
flow rates and compute provisional
  calibration slope and intercept
         (Steps 23. 24)
                   %dffl. within ±6%
          Audit flow rate %
          difference is firm
                                                          Remove orifice; operator measures
                                                            stagnation pressure at normal
                                                            operating flow rate, with fitter
                                                                 (Steps 25. 26. 27)
  Remove orifice; operator measures
    stagnation pressure at normal
     operating flow rate, with filter
          (Steps 16.17. 13)
                JL
                                                                       JL
                                                        Calculate Qa (sampler) and Qa (audit)
                                                            at normal operating flow rate
                                                                    (Step 28)
Operator determines indicated flow rate,
 Qa (sampler), at normal operating flow
           rate. (Step 19)
                                                                       I
                                                         Compute audit flow rate % difference
                                                           of Qa (sampler) with respect to
                                                              Qa (audit) (Steps 29,30)
  Calculated Qa (corrected sampler)
   using audit flow rate % difference
             (Step 20)
                                                        Compute design flow rate % difference
                                                        of Qa (audit) with respect to design flow-
                                                                rate. (Steps 31.32)
     Calculate design flow rate %
difference using Qa (corrected sampler)
           (Steps 21. 22)
                t
 One-Point Audit is Complete
                                                                       t
                                                        Three-Point Audit is Complete
              Figure 7.2. Flow chart of performance audit procedures for
                       volumetric-flow-controlled PM10 samplers.

-------
                                                            Section No.: 2.11.7
                                                            Date: January  1990
                                                            Page: 9


  1.   Transport the following equipment to the monitoring site:

           Audit orifice transfer standard wi.th calibration relationship in
           actual volumetric flow-rate (Qa) units and traceable to NIST.  The
           audit orifice transfer standard's faceplate or filter cassette may
           require modification to ensure a good seal during the performance
           audit.  The audit orifice transfer standard should not be the same
           one that is used for routine calibrations and QC flow checks.

      •    An associated water or oil manometer,  with .a 0- to 400-mrn (0- to
           16-in.)  range and minimum scale division of 2 ram (0.1 in'.).

           A thermometer,  capable of accurately measuring temperature  over an
           appropriate range to the nearest ±1  «C and referenced to  an  NIST or
           ASTM thermometer within ±2 °C  at least annually.

           A portable  aneroid barometer (e.g.,  a  climber's or  engineer's altim-
           eterj, capable  of accurately measuring ambient barometric pressure
           over the  range  of approximately 500  to 800 mm Hg  (66  to 106  kPa)  to
           the nearest mm  Hg and  referenced within  ±5 mm Hg  of a barometer  of
           known  accuracy  at least  annually.,

  •    •     VFC Sampler Audit Data  Sheet such as shown in  Figure  7.3  (blank
           forms  appear in  Subsection  12).

      •     A  clean filter.
            The S1te operator is respons-.ble for providing the manometer  (or
     <™nP^  Te m?aSUring device) that is n°™a11y used for measuring the
     sampler s stagnation pressure, the sampler lookup table or alternative
     ffow r^I°^ecatTShip ^al is current]y in effect for determining the
     thS? i* ™™ i?  P1ln.9 Periods,  and any other information or equipment
     that is normally used to determine the sampler's indicated flow rate.

2.   Instruct the operator to install a clean filter in the VFC sampler.  A
     nr™!riCa5   iVay be US!d if the audit orifl'ce transfer standard can be
     properly mounted on top of the cassette. -Otherwise,  install  the filter
     without the cassette.  An audit  filter should never be used for subs"  '
     thTfilS  h?i   thUSe Partic1es.larger than '10 jm can be collected on
 '    a  rplu   f    * the inlet is  raised.-  The sample mass will, be biased as
     a  result  of using a filter for both an  audit  and subsequent sampling

3'   Check'1 Jh^%hpd^ ?rJfl'Ce transfe!: standard's  faceplate on the sampler..
     Check  that the  gaskets  are in  good condition  and have not deteriorated.

     Caution:   Tighten  the faceplate  nuts  evenly on  alternate  corners  to

     ?ia°hPt±Ha1T  T1  unif«r»ly S6at  the gask*tS'   The nut*  sh™d be hand-
     tightened  only;  too  much  compression  can damage  the sealing gaskets

-------
                                                                            Section No.:  2.11.7
                                                                            Date:  January 1990
                                                                            Page:  10
                          VFC SAMPLER AUDIT DATA SHEET (Part 1)
Station Location  -*-/*" **' *™ t roL-\^>    Date     /  I-2-^1 V I Time
Sampler Model    W.gJ>£ //Vg-	 s/N     /OV & ~7	;,       Orifice Calibration Date   (~2-\ I S~ I &  o
Orifice Calibration Relationship:  m «   t'l-^go    D _ -a.ooCa,  r ,   Q
Sampler Calibration Relationship:  m -   L^qolfoP  b  -   7~A&£-£  r -
Orifice Pressure Drop UH2O)  */-*?"   in. H2O     Qa (audit)3    f' ° ^ ^> _ m3/min
                                      With Orifice Installed         Without Orifice Installed
Stagnation Pressure UPstg)          _ -3 1 * "3> "7    mm Hg        3 3. o 2 —   mm Hg
Absolute Stagnation Pressure (P1)5       £ *7 *?• & -3    mm Hg     "7 ^ .£"* 7 ^    mm Hg
Stagnation Pressure Ratio (P1/Pa)           O fe?¥6~7  (see Note)       ^-^ ^V-3 (see NOte)
Qa (sampler)0  '                          /- / 2.  /      m3/min       / ^/V  ^      m3/mjn
Audit Flow Rate Percentage Difference*1    ^^ ^    "/o (see Note)
Qa (corrected sampler)"  _ t'flf~^> _ m3/min
                                   ,           tJ. -)
Design Flow Rale Percentage Difference'  _ '    *• ^ — _ %
Note: If Pi/Pa is less than the values that are listed in the manufacturer's lookup table or if audit flow rate
      percentage difference is greater than ± 6 percent, proceed to Part 2 of the VFC Sampler Audit
      Data Sheet. Otherwise, complete this part.
    (orifice) - {U(H2O)(Ta/Pa)r/» - b}  {1/m}
bP1 - Pa - .iPstg
cDetermine Qa (sampler) from manufacturer's lookup table (or from alternate calibration relationship)
dAudit % Difference »   (Qa (sampler) - Qa (audit)
                              Qa (audit)
                                                [100] where Qa (sample) is measured with the
                                                      orifice installed.
eQa (corrected sampler) » Qa (sampler)
                                     100 - Audit % Difference
                            •[•
 _  .   _,   _   „, _.„          Qa (corrected sampler) - 1.13
'Design Row Rate % Difference
                                                              where Qa (sampler) is measured
                                               100             without the orifice installed
                                            1-13
Auditor   R. ^ I a- ft T- _        Observer    K •
                                                                                (continued)
                Figure 7.3.  Example VFC sampler audit data sheet. (Part 1 of 2)

-------
                                                                           Section  No.:  2.11.7
                                                                           Date:  January  1990
                                                                           Page;  11
                            VFC SAMPLER AUDIT DATA SHEET (Part 2)
                                (continued from Part 1 of this form)
  Station Location  /VJP/JA/A Pe Lf 3	 Date      '
  Sampler Model —W£T>Ki AS/Z.	 S/N    ( g V^>7^
  ** -2-^2—mmHg, Ta 	(_L	°C  _2rfL^K, Unusual Conditions
  Audit Orifice S/N	
                                                                        Time  3-," /£
  Orifice Calibration Relationship:  m -  A
                                               Orifice Calibration Date
                                                    b -  —
  Provisional Sampler Calibration Relationship: X = Qa (Orifice), Y * (Pi/Pa)
                     ~
                                                                              Qa (Orifice)'
                                                                               (m3/min)
Measurements with filter installed and Audit Orifice Transfer Standard removed-
APst9  -   33.01-  • - _ mm Hg
Pi/Pa _ <^ ,  <
Qa (audits      /
                                      m3/mjn     Qa (samp,er)d
Audit Flow Rate Percentage Difference6 _
Design Flow Rate Percentage Difference'
                                               -V. a
 aP1 «. Pa - APstg
 bQa (orifice) . {[(AH2O) (Ta/Pa)J1/z - b}  {1/m}
 cQa (audit) « {(Pi/Pa) - b'|/m'
 dOa (sampler) is determined from the lookup table or .alternative calibration relationship
eAudit Flow Rate o/o Difference = RQa ^rnp'er)  - Qa (audit)]
                              I       Qa (audit) .       j L100J

'Design Flow Rate % Difference - [ Qa (audit) ~ 1-131 [100]
                               L       1.1*3      j
Auditor
                                                                                .o
.mm Hg

. m3/min
                Figure 7.3. Example VFC sampler audit data sheet. (Part 2 of 2)

-------
                                                           Section No.: 2.11.7
                                                           Date: January 1990
                                                           Page: 12


4.   Install the audit orifice transfer standard with no resistance plate,  or
     open the valve of a variable-resistance orifice wide open.  For resist-
     ance plate-type orifices, make sure the orifice gasket is present and
     that the audit orifice transfer standard is not cross-threaded on the
     faceplate.  Seal the audit orifice transfer standard's pressure port and
     the stagnation pressure port with rubber caps or similar devices,

5.   Leak test the audit system-(refer to Subsection 2.4.2, Step 4).  Identify
     and correct any leaks before proceeding.

6.   Inspect the audit manometer connecting tubing for crimps or cracks.
     Fully open the valves and blow gently through the tubing, watching for
     the free flow of the fluid.  Adjust the manometer sliding scale so that
     the zero line is at the bottom of the meniscuses.  Connect the audit
     manometer to the pressure port on the audit orifice transfer standard.
     Make sure the unconnected side of the manometer is open to the atmos-
     phere.  Make sure that the tubing fits snugly on the pressure port and on
     the manometer.

7.   Read and record the following parameters on the VFC Sampler Audit Data
     Sheet:

          Sampler location, date,  time.
          Sampler model and S/N.
          Ambient temperature (la),  °C and K (K = °C + 273).  .
          Ambient barometric pressure (Pa), mm Hg (or kPa.).
          Unusual weather conditions.
          Audit orifice transfer standard's S/N and calibration relationship.
          Sampler lookup table number or other calibration relationship cur-
          rently in effect.

8.   Turn on the sampler and allow it to warm up to operating temperature (3
     to 5 rain).

     Note;  The sampler inlet may be partially lowered over the audit orifice
     transfer standard to act as a draft shield (if a shield is not otherwise
     provided).  Use a block to provide at least 2 in. of clearance at the
     bottom for air flow and for the manometer tubing.

9.   When the sampler has warmed up  to operating temperature,  observe the
     pressure drop across the orifice by reading the total  manometer deflec-
     tion and record as AH2<3 on the  audit data sheet.

10.  Instruct the operator to measure the sampler's relative stagnation pres-
     sure (i.e. relative to atmospheric pressure)  with the manometer (or other
     pressure measurement instrument)  normally used to measure stagnation
     pressure.   Record the relative  stagnation pressure as  APstg on the data
     sheet.  If APstg is measured  in inches of water, convert the reading to
     mm Hg using Equation 12:

-------
                                                            Section  No.: 2.11.7
                                                            Date: January  1990
                                                            Page: 13


                         mm Hg = 25.4 (in. H20)/13.6                   (Eq. 12)

 11.  Compute the absolute stagnation pressure, PI, as:

                               PI - Pa - APstg                         (Eq. 13)

      and the absolute stagnation pressure ratio as:

                      Stagnation  pressure ratio = Pi/Pa                (Eq. 14)

      Record  the  Pi/Pa ratio  on the audit data sheet.
                 Qa(audit) -  {[4H20(Ta/Pa)]l/2  -  b}  {1/n}               (Eq.  6)
     where:

          Qa(audit) = actual  volumetric flow rate as indicated by the audit
                      orifice transfer standard, m3/min  '
               AH20 = pressure drop across- the orifice, mm  (or in.) H?0
                 Ta = ambient temperature, K (K = °C + 273)
                 Pa = ambient barometric pressure, mm Hg (or kPa)
                  b = intercept of the audit orifice transfer standard's cali-
                      bration relationship                   .
                  m = slope of the audit orifice transfer standard's calibra-
                      tion relationship.

13"       rUC  tl'e °P
-------
                                                           Section  No.: 2.11.7
                                                           Date: January  1990
                                                           Page: 14


16.  Turn  off  the  sampler  and  remove  the  audit orifice transfer  standard.

17.  With  a  filter installed on  the sampler  in the normal sampling  configura-
     tion  (i.e., with  a  filter cassette,  if  normally used), turn on the sam-
     pler  and  allow it to  warm up to  operating temperature.

18.  Instruct  the  operator to  measure the sampler's relative stagnation pres-
     sure, APstg,  and  to calculate the absolute stagnation pressure ratio,
     Pi/Pa,  as specified in Steps 10  and  11.  Record these data on  Part 1 of
     the audit data sheet.  Turn off  the  sampler, and replace the vacuum cap on
     the stagnation pressure port.

19.  Calculate the sampler's indicated operational flow rate, Qa(sampler),
     using the Pi/Pa value obtained in Step  18 and the sampler's lookup table
     or alternative calibration  relationship.  Record this flow rate on Part 1
     of the  audit  data sheet.

20.  Calculate the corrected sampler  flow rate using Equation 23:

     Qa(corrected  sampler) = [Qa(sampler)] [10° " Audi^ Difference] (Eq_ 23)


     where Qa(sampler) is  obtained from Step 19 and the audit flow rate per-
     centage difference  is obtained from Step 14.  Record this value on Part 1
     of the  audit  data sheet.

21.  Calculate the  design  flow rate percentage difference between the cor-
     rected  sampler flow rate.from Step 20 and the inlet design flow rate of
     1.13 nH/min as:

     Design  flow rate  _   FQa(corrected sampler)  - 1.131  finnl         /c   1QN
     % difference      ~   I            1713	J  [100j         (Eq" 18)


     Record this value on  Part 1'of the audit data sheet.

22.  If the design  flow  rate percentage difference is within ±7 percent, the
     sampler calibration is acceptable.  Differences exceeding ±7 percent
     should be investigated.  Differences exceeding ±10 percent (or the
     acceptable flow-rate  range specified by the inlet manufacturer) may
     result in the  invalidation of all data obtained subsequent to the last
     calibration or  valid  flow check.   Before invalidating any data, double-
     check the audit orifice transfer standard's  certification,  and all  cal-'"
     culations.

     This completes  the one-point audit.   Return  the sampler to its normal
     operating configuration.

23.  Repeat Steps 8  through 12  for at  least two additional  audit flow rates,
     obtained with  resistance plates  or by adjusting the  variable resistance        ^

-------
                                                           Section No.: 2.11.7
                                                           Date: January 1990
                                                           Page: 15


     valve.  These additional flow rates should be within or only sliqhtlv
     below the acceptable flow-rate range (i.e., 1.02 to 1.24 m3/min) of the
     sampler inlet, if possible.  Record the data from all three points on
     Part 2 of the audit data sheet.

     Note:  Resistance plates supplied by the manufacturer of the audit ori-

     IJSftJnJ!!? JriSt?!ld-rt !?3y Hav! t0 be modified (i.e., holes enlarged or
     additional holes drilled)  to obtain appropriate flow rates.  -•

24.  For each calibration- point, calculate and record the quantities  (Pi/Pa)
     and Qa(onfice).   For the  linear regression model,

                      (Pi/Pa) = m'[Qa(orifice)]  + b1                   (Eq.  19)

     Calculate  the linear regression  slope  (m1),  intercept (b')  and correla-
     tion coefficient  (r«) .   Be sure  to  include  the  data  obtained for [he
     rinJl  -°W ^TK  in  StePs.n and  12  in the calculation  of the linear
     regression.   The  regression calculations  require  a total of at least
     three points.  Record  these values  on Part  2  of  the  audit data sheet.

     £pte:  This  is a  temporary,  provisional calibration  relationship  that  is

     thVca°cuTla9            be°aUSe n° temPerature te™ is  included in
      thVcacuTlations

25.   Turn off the sampler and remove- the audit orifice transfer standard.
26'  ?ion  r/i1161"-^513]-?? °n the Sampler in the normal sampling configura
     lllr  «d inn V*   tCr cassette' if normally used), turn on the lam-
     pier  and allow it to warm up to operating temperature.
     sureAPst     Htnr                5fP1er'5 r*]ative stagnation pres
     Pl/Pa « Itrif  J°.ca^ulate the absolute stagnation pressure ratio,
                         n S6S 10 and n«   Record these data °n Part 2 of
28'  SjISn1?!6 DwDSamp!er'S indicated operational  flow rate,  Qa(sampler)
     or altern.t  7   V? H* ?taine? 1n Step 27 and  the sampler's lookup  able
     ?L IISKr  ve ca1l5ratlon/elationshiP.   Calculate the audit value for
     the sampler's operatTonal  flow rate,  Qa(audit),  using  the (Pi/Pa)  value

     ?i^?1lniStep ?7Kand the Slope (m')  and intercePt9(b')  if the prb   -
     sional  samp  er calibration  relationship obtained  in Step  24   Record
     these two flow rates  on Part  2 of the audit  data  sheet
                                                        [IOQ]

-------
                                                                Section No.: 2.11.7
                                                                Date: January 1990
                                                                Page: 16
          Record this value on Part 2 of the audit data sheet.
     30.  If the audit flow rate percentage difference is within ±7 percent, the
          sampler calibration is acceptable.  Differences exceeding ±7 percent will
          require recall" brat ion.  Differences exceeding ±10 percent may result in
          the.invalidation of all data subsequent to the last calibration or valid
          flow check.  Before invalidating any data, double check the sampler's
          calibration, the audit orifice transfer standard's certification, and all
          calculations.

     31.  Calculate the design flow rate percentage difference between the
          Qa(audit) obtained from Step 28 and the inlet design flow rate of 1.13
          m-Ymin as:

               Design flow rate  _  fQa(audit) - 1.131 finn]                /P   ,«>
               % difference      ~  I       1.T3"     J 1100J                (Eq' 18)

          Record this value on Part 2 of the audit data sheet.

     32.  If the design flow rate percentage difference is within ±7 percent, the
          sampler calibration is'acceptable.  Differences exceeding ±7 percent
          should be investigated.  Differences exceeding ±10 percent (or the
          acceptable flow-rate range specified by the inlet manufacturer) may
        •- result in the invalidation of all data obtained subsequent to the last
          calibration or valid flow check.  Before invalidating any data, do.uble
          check the audit orifice transfer standard's certification, and all calcu-
          lations.

          This completes the three-point audit.  Return the sampler to its normal
          operating configuration.

7.3  Audit Data Reporting -

     The operating agency should be given a copy of the audit preliminary results
at the completion of the audit.   The audit data sheet should be signed by.both the
auditor and the observer, and the results should be discussed.  These preliminary
data should never be used to make monitoring system adjustments.  Auditors may make
mistakes, and calibration curves may shift.  A post-audit verification of audit
equipment and data is essential  before inferences can be drawn regarding the samp-
ler's performance.  An auditor should be able to "support audit data with complete
pre- or post-audit equipment verification documentation.

     Final verified audit data should be submitted to the operating agency as soon
as possible.  Delays may result  in data loss;  a sampler out of audit limits is also
out of calibration limits,  and the data collected may.be invalid.  If a sampler
exhibits unsatisfactory agreement with the verified audit results (i.e., audit flow
rate percentage differences exceeding ±7 percent),  a calibration should be per-
formed immediately, certainly before the next run day.

-------
                                                                   Section No.: 2.11.7
                                                                   Date: January 1990
                                                                   Page: 17
   7-4   Flow-Rate  Performance Audit Frequency -

  7.5  System Audit -
                                               additional detailed procedures and
      1-.   Use the operational  flow rates as reported  on  the  sample  data  sheets

      "
ance operator's technique  the auditor^hnnirt'rJl! ajd1ti°n *« observing  the  bal-
posed filter reweighinj as discussed "n Sect?™ ? ¥l *1 ^de^endent  ™<^  of ex-
formance evaluation of the HV PM10 fiitPr w«!nJ-       °   his  volume-   "The  per-
the following manner:                    r wei9»"ng process should  be  conducted in

-------
                                                           Section No.:  2.11.7
                                                           Date: January 1990
                                                           Page: 18


1.   Instruct the operator to weigh a set of three to five Class S standard
     weights covering the range normally encountered during gravimetric  analy-
     sis.  The weighed value should agree within ±0.5 mg of the stated values.

2.   Observe the balance operator's technique, and review the laboratory
     weighing procedure for determining both the tare and gross weights  of the
     sampling filters.

3.   Review the maintenance and calibration log for each balance.  Routine
     balance maintenance and calibrations must be performed by the manufactur-
     er's service representative at manufacturer-specified scheduled inter-
     vals.  In no case should the interval between calibrations exceed 1 year.

A.   Review QC data records for the filter-weighing process.  Ensure that the
     following QC activities have been performed and documented:

     •  .  Standard weight check every day of balance operation.

     •    Zero checks after every 5 to 10 filter weighings, calibration  checks
          after every 15, unless longer term stability of the balance has been
          established by records (at least 1 calibration check per day).

     •    At least five filters reweighed each day.  '

     If any of these QC checks are out.of limits, note what action was taken.

-------
Section No.: 2.11.7
Date: January 1990
Page: 19









10
t-
EQUIREMEN
cc
AUDITING
UJ
o
i-










i
j
i












1
o
c
u
c •
o
u c
•"{
o-
L.
w i
CX 49
•- 0 1 « *
E -x v. « • o
3 0 U 0 K — —
« U 0 «. S «•-
0 C X «- « 0
U 0 U — tl O "D
t. 1 -0 c —
000 !!• 	
u v- — •*. ^ n •
O «•- -O >-( 0 >
«•- •- 3 0 •
• "DO O 49 1.
^ -a • x • o
*• « 0 -o
0 — - c c
4> K 0 — 0 0
i. d r* 49 .- • 49
-O C »l • 49 -O •
— •- — « — L. .
— -^_» 3 — .0 J«
o|0-*>;i0
O:»ou0— o"u
1 -o
0 C

49 .—
• #»
i U 1 ~

'I
eptance 1
u
u
TJ
O
'.c

U
O
-o
c

x
u
c
0
3


U.



0
L.
3
-o
e
u
0
a.

0 .- 49 .-
Pi *
c , ...
*" * * £
— a 49 •
•06 u
• -o
0 « C (
49 X II C
C-49 0
0 49 -O
U C • 0
t- • t_ L.
00 3
a. s * •
49 0 • •
• •- fK
«-> .O «fc £ (••
c <
cvc
en o
a. s-
L. L.
O •
*. 0
0 1.
49 .0
L 0.
3 0
er o
C
f Q

£ K
0 ca
c
e —
u v.
C 0
O 49

49

T3
3

49

1
O
U.
,
•• t_
st flow control ler
ary (MFC sampler) 01
t manufacturer (VFC
r) . Recalibrate.
3 « U 0
T— 1 II • —
•o 0 49 a
• u c E
0-0 0 •
a: c u •

y
O
^
01
T>
C
0
^p
P4 •


























te improved methods
training programs.
• L.
— -o
*~ ^9
C C

*2
.
Q
!c c
.c'|
(escribed
on and Sa
Handbook.
W "^
-o o —
O 0 X
1 • 0
L. • U
0 3
49 XTS
"S- 0
§• O
u u
— Q.
2??
L. L •
e 0 0 49
a. » »
"S-o-gz
o • • u
e

e * e u
e 49 —

0 • O 3 -
•O IV
4> e a"c




**
•o
3
e|

j
M
(XI


kail calculations.
u
0
ji
u
2


\
c
• 0
0 U
0 C «.
I- O •*•
01 O O
• 1
49-0
cue
O O 3
— a o
49 0 t
e u
i. e
+} «^ •«
c • jr.
0 C 49
e'a'J
o •—
u u c
00-
49 ™ U
— X 49 O
T3 49 • u
3 — L L
< * 49 0
• .-.
t 0 49
•3d-
0 O • L.
•S3"!
o a r»
• e o v.
0 e «- o
a. u
0 c -o E
u o u 5
u o §
• x o —
- ® 0 e
49 »•» u •—

0 CL> e ^^
^ 4>
e *• • «B
• O -D S

•o o o k'
53ii


a»
c
•^
•1
e
u
o
a.
s
Q


0
49
c
I)
I.
e
o>
* 49



•o
n
— o
• CL
C 'X
L
O U.
O
It •»•
c ^1 *
49 X 0
— OI49
• e
e 0 o
t- « —
2 3.-.c
— • c ^
*• t. 0 CM
0 U
B 49 49
& — •- •
•2s S ~ a
.— U> — 49 C
•a vi • e •—
• *« "^


f^ •.• • 4B. A
^ U ft

>- • 49 k 0
0 X 0 I.
«•• ^ — 49 O
t. • — «.
0 1. C ~ 0
a. o e «. jo


a>
e

a
0
*
u
0
IT

1
i
0
*
u
49
U
— e
— e
X ^3
d

•
X
^
•e •
e 3
• c
c
•?-
C 49
•• •
e •
I-
• 49
E •
• -a
•— 0
49
0 •
u t-
c -n
• —
•e •
n u
c
S8
e- c
c c
X 0
U 49
0 e
e a
— 0 0
^ t* —
a c


• ^
* 5 "^

> * u

a u u








u
c
0

-------

-------
                                                      Section No.: 2.11.8
                                                      Date: January 1990
                                                      Page: 1
2.11.8  ASSESSMENT OF MONITORING DATA FOR PRECISION AND ACCURACY

8.1  Precision

fn  .One.or\more monitoring sites within the reporting organization are selected
for duplicate, collocated sampling as follows: "for a network of 1 to 5 sites  1
sue is  selected; for a network of 6 to 20 sites, 2 sites are selected- and for

         f ^ i -han-2° SlteS' 3 SiteS are Selecied-  Wherfpo s ble
 cc       i --
 of S «. fS11"!  1S ?nC?!Ta9ed- Annual mean P^ticulate  atter concentrons
                    h     F "9 ^he hl'9hest ^ Percent of the annual mean

8.2  Accuracy
ffisL«HE"SiF:vF;r-': -"•™™«« ™-':«-

-------
                                                                   Section No.:  2.11.8

                                                                   Date: January 1990

                                                                   Page: 2
$2


LU

CC
Q.
 s
 cc
 O
 u.

 a
 z
 H-
 
•^ u ?!
z
00
> <*•
< ri




a
Q UJ O
X Q M
>- O -
2
is


UJ
° ° aj
O (j T
= uj 0
5 = ""






























«•








































»•

















































































o-

















































































a.


























































































































a.



































































































































































«-








































«•

















































































a.








































a.



































































































































































0,

















































































a.
















1























0.















-------
 5
 cc
 O
 cc
 O
 Q.
 LU
 CC
 CO


 CO
 CO
0 1
j* ?
< 3
Q -,
                                                                       Section No.:  2.11.8
                                                                       Date:.January 1990
                                                                       Page: 3
  a
 o     S  P'
 2     2  tj-
  z

 II
 02
, «.<
 sg
z
o
M
<
•*
Z

o
c
a
a
z
        M     *  i
               •!  S
          «.    *  2
** •••
«
•• "™"

tk ••
! 2
, 3 S
< ^





9

Z 0



a
i
M
       a
       0
       u
VI
O
VI
1'

a
u
1 POUUIANI


-------

-------
2.11.9  RECOMMENDED STANDARDS FOR ESTABLISHING TRACEABILITY
                                                                 Section  No.: 2.11.9
                                                                 Date: January  1990
                                                                 Page: 1
n,nr     **???* *? essential for attaining accurate data:  (1) the measurement
process must be under statistical control at the time of the measurement, and
(2) the combination of systematic errors and random variation (i.e., measurement
errors) must yield a suitably small uncertainty..  The attainment -of accurate data
requires the performance of QC checks, independent audits of the flow measurement
process-  careful documentation of monitoring data, and the use of equipment and
standards that can be traced to appropriate primary 'standards.

9 '• 1 . Recommended Standards for Establishing Traceabilitv

     1.   Class-S weights of NIST specifications are  recommended for the laboratory
          balance calibration.   See Subsection  4.5 for details  on balance cali-
          bration checks.

     2.   Use of a positive-displacement  standard volume meter  (e.g.   a RootsR
          Meter)  is recommended  for calibrating the flow-rate transfer standards
          that  are used  to calibrate  and  audit  the HV  PM10  sampler.

          Note:   As they  are sold,  standard  volume meters may not  be  traceable  to
          NIST    Traceability can  be  established  through  NIST or through  the meter
         manufacturer's  repair department.   Periodic  recertification  is  not nor-
     "   !fl y  Iequyed  "nder Clean  servi«  conditions  unless the meter  has been '
         damaged  and must  be repaired.   Subsection 2  presents, detai Is  on HV PM10
         sampler calibration, and  Subsection 7 presents details on the flow-rate
         performance audits.
    3'   Inr J;13?5?^"16 met^ Sh°u1d be checked UP<™ initial receipt and refer-
         min/da         an™ally against an accurate timepiece to within 15


    4.   The accuracy of associated monitoring equipment (i.e., thermometers

             nffL5' ,SJ°P yVatCheS'  etC')  Sh°uld be check^ at routfnTintervals,
                                            '
          un                                                                 ,
         traceable to NIST0"06 Per y6ar' a9ai'n5t 5tandards of known accuracy and

-------

-------

-------

-------