EPA600/
2-84-187                                                       PB85-122737
           RECOMMENDED PRACTICE FOR THE USE OF ELECTROMAGNETIC
           FLOWMETERS IN WASTEWATER TREATMENT PLANTS
           National Bureau of Standards
           Washington, D. C.
           NOV 84
                            U.S. DEPARTMENT OF COMMERCE
                          National Technical Information Service

-------

-------
                                             EPA-600/2-84-187
                                             November  1984
RECOMMENDED PRACTICE FOR THE USE OF ELECTROMAGNETIC
     FLOWMETERS IN WASTEWATER TREATMENT PLANTS
                        by

                   Gershon Kulin
            Fluid Ennineerinf Division
           National Bureau of Standards
              Washington, D. C. 20234
                 EPA 78-D-X0024-1
                  Project Officer

                  Walter W. Schuk
           Wastewater Research Division
    Municipal  Environmental Research Laboratory
              Cincinnati, Ohio 45268
   MUNICIPAL  ENVIRONMENTAL RESEARCH LABORATORY
        OFFICE OF RESEARCH AND DEVELOPMENT
       U.  S. ENVIRONMENTAL PROTECTION AGENCY
              CINCINNATI,  OHIO  45268
               REPRODUCED BY
               NATIONAL TECHNICAL
               INFORMATION SERVICE
                   U.S. DEPARTMENT OF COMMERCE
                     SPRINGFIEID, VA. 22161

-------

-------
                           (flease read Instructions on the reverse before completing)
1. REPORT NO.

  EPA-600/2-S4-137
4. TITLE AND SUBTITLE
    RECOMMENDED PRACTICE  FOR  THE  USE OF ELECTROMAGNETIC
    FLOWMETERS  IN WASTEWATER  TREATMENT PLANTS
     5. REPORT DATE
      November 1984
     6. PERFORMING ORGANIZATION CODE
               s ACf §s
ON NO.
 737
 AUTHOR(S)
    Gershon Kulin
                                                          8. PERFORMING ORGANIZATION REPORT NO.
 PERFORMING ORGANIZATION NAME AND ADDRESS

    National Bureau of  Standards
    Fluid Engineering Division
    Washington, DC  20234
      10. PROGRAM ELEMENT NO.

          B113, CAZB1B
      11. CONTRACT/GRANT NO.
          IAG  No.  EPA-78-D-X0024-1
12. SPONSORING AGENCY NAME AND ADDRESS
    Municipal Environmental  Research Laboratory--Cin., Of
    Office of Research and  Development
    U.S. Environmental Protection  Agency
    Cincinnati, Ohio  45268__....	
      13. TYPE OF REPORT AND PERIOD COVERED
          Handbook--!0/1 /78-9/30/81
      14. SPONSORING AGENCY CODE
          EPA/600/14
15. SUPPLEMENTARY NOTES
    Project Officer:   Walter  W.  Schuk     Telephone -  (513)  684-2621
16. ABSTRACT"

         Electromagnetic  flowmeters  that conform to the guidelines  described in this
    document can be used  to measure  the volumetric flowrate of  all  liquids and sludges
    normally encountered  in wastewater treatment plants, provided that  adequate
    inspection and maintenance  are  performed as recommended to  contend  with potential
    deposits and other effects  of the harsh fluids.

         Electromagnetic  flowmeters  should be accurate to the lesser  of 1  percent of
    full-scale or 3 percent of  actual  flowrate.  These flowmeters are not  necessarily
    immune to the effects  of  approach velocity distribution.  Specific  approach
    conditions are cited  which  must  be observed during installation in  order to
    minimize velocity-distribution  errors.

         Upon installation, electromagnetic flowmeters should be hydraulically tested,
    by a method equivalent to that  described, to provide an acceptance  test and/or
    initial calibration.   Continuing and frequent performance monitoring subsequent to
    initial testing is particularly  important for electromagnetic flowmeters  in treat-
    ment plants.  Methods  for accomplishing this are suggested.
                                KEY WORDS AND DOCUMENT ANALYSIS
                  DESCRIPTORS
                                              b.IDENTIFIERS/OPEN LNDED TERMS
                      COSATI Reid/Group
18. DISTRIBUTION STATEMENT

       RELEASE TO PUBLIC
                                               9 'SECURI !"Y CLASS / I Ilix kfp.irn     21. NO OF PAGES
UNCLASSIFIED
     39
                                                     UNCLASSIFIED
 EPA Form 1220-; ;R«

-------
                                 DISCLAIMER
       Although the information described in this document has been funded
wholly or in part by the United States Environmental  Protection Agency
through assistance agreement number EPA 78-D-Xti024-l  to National  Bureau of
Standards, it has not been subjected to the Agency's  required peer and
administrative review arid therefore does not necessarily reflect the views
of the Agency and no official endorsement should be inferred.

-------
                                   FOREWORD

     The U. S. Environmental Protection Agency was created because of in-
creasing public and Government concern about the dangers of pollution to the
health and welfare of the American people.  Noxious air, foul water, and
spoiled land are tragic testimonies to the deterioration of our natural
environment.  The complexity of that environment and the interplay of its
components require a concentrated and integrated attack on the problem.

     Research and development is that necessary first step in problem solu-
tion; it involves defining the problem, measuring its impact, and searching
for solutions.  The Municipal Environmental Research Laboratory develops new
and improved technology and systems to prevent, treat, and manage wastewater
and solid and hazardous waste pollutant discharges from municipal and communi-
ty sources, to preserve and treat public drinking water supplies, and to mini-
mize the adverse economic, social, health, and aesthetic effects of pollution,
This publication is one of the products of that research and provides a  most
vital communications link between the researcher and the user community.
                                     m

-------
                                   ABSTRACT

     Electromagnetic flowmeters that conform to the guidelines described in
this document can be used to measure the volumetric flowrate of all liquids
and sludges normally encountered in wastewater treatment plants, provided that
adequate inspection and maintenance are performed as recommended to contend
with potential deposits and other effects of the harsh fluids.

     Electromagnetic flowmeters should be accurate to the lesser of 1  percent
of full scale or 3 percent of actual flowrate.  These flowmeters are not nec-
essarily immune to the effects of approach velocity distribution.  Specific
approach conditions are cited which must be observed during installation in
order to minimize velocity-distribution errors.

     Upon installation, electromagnetic flowmeters should be hydraulically
tested, by a method equivalent to that described, to provide an acceptance
test and/or initial calibration.  Continuing and frequent performance  moni-
toring subsequent to initial testing is particularly important for electro-
magnetic flowmeters in treatment plants.  Methods for accomplishing this are
suggested.
                                      iv

-------
                                  CONTENTS

Foreword 	 i i i
Abstract 	  iv

   1.  Scope 	   1
   2.  Definitions 	   2
   3.  Background 	   3
       3.1   Principle of Operation 	   3
       3.2  Advantages and Disadvantages of Electromagnetic
            Flowmeters 	   4
   4.  Guidelines for Specifications 	   6
       4.1   Flow Tube 	   6
       4.2  The Secondary 	   8
       4.3  Calibration 	   9
       4.4  Accuracy 	   9
   5.  Installation Requirements 	  10
       5.1   Approach Conditions 	  10
       5.2  Bypass 	  10
       5.3  Orientation and Location 	  10
       5.4  Limiting Velocities 	  11
       5.5  Other Considerations 	  11
   6.  Error Sources 	  13
       6.1  General 	  13
       6.2  Velocity Profile 	  13
       6.3  Deposits 	  14
       6.4  Air or Other Gases  	  15
       6.5  Zero Drift 	  15
   7.  Calibrations and Performance Checks 	  16
       7.1  Hydraulic Calibration of the Measuring System 	  16
       7.2  Calibration Other than Hydraulic	  19
       7.3  Flow Measurement Methods for Hydraulic Calibrations ....  20
       7.4  Flow Measurements for Monitoring 	  25
       7.5  Approximate Flow Measurements 	  27
   8.  Operati on and Mai ntenance 	  30
       8.1   Operation/Maintenance Frequencies 	  30
       8.2  Miscellaneous Operation/Maintenance Recommendations ....  31
   9.  References 	  32

-------

-------
                                   1.   SCOPE

1.1   This document covers the use of electromagnetic  flowmeters  for in-plant
     flowrate measurement of raw influent,  treated  effluent,  and wastewater  in
     intermediate stages of treatment as  well  as  various  types of liquid  sludges,

1.2  For the purposes of this document, "use"  of  electromagnetic flowmeters  is
     considered to include:  requirements  for the  design and construction  of  the
     primary and secondary elements;  requirements for installation of the equip-
     ment; recommendations for performance  checks,  operation  and maintenance.

-------
                                2.  DEFINITIONS

2.1  Accuracy -- The sum of the systematic error (bias or difference from the
                 "true" value) and the precision at a stated confidence level
                 (e.g., in terms of a number of standard deviations).  In the
                 case of flowmeters, the true value is usually determined from
                 a basic calibration method.

2.2  Flow converter — The element of the secondary system that converts the
                 output electrical signal to flowrate units on a dial, digital,
                 chart or other readout.

2.3  Flow tube -- The part of the primary that forms a pipe-like conduit.

2.4  Precision — The reproducibility or repeatability of a measurement.

2.5  Primary -- The element of a measuring system that generates a measurable
                change in a parameter.  In this case, the primary consists of
                the flow tube, the coils and yoke that produce a magnetic flux,
                and the electrodes.

2.6  Quadrature noise -- Electrical noise that is 90 degrees out of phase with
                the desired flow signal.

2.7  Repeatability -- See precision.

2.8  Reynolds number -- A dimensionless number that expresses the ratio between
                inertial and viscous effects in a flow.   Low Reynolds numbers
                (below about 2000 for pipe flow) correspond to laminar flow.

2.9  Secondary -- The portion of the measuring system that amplifies and/or
                conditions the signal from the electrodes into an output pro-
                portional to the flowrate and further usually converts it to
                a readout in flowrate units.  Part or all of the secondary
                may be mounted directly on the primary.

2.10 Threshold conductivity — A minimum electrical conductivity above which
                changes of conductivity of the flowing liquid have a negligi-
                ble effect on the measurement.

2.11  Weight function -- A theoretical function that weights the contribution of
                each part of the induced voltage in the  flow section to the
                total  sensed by the electrodes.

2.12 Zero drift -- A change with time in the output reading for the no-flow
                condition.

-------
                                3.   BACKGROUND
3.1  Principle of Operation
     3.1.1  Flow-tube meters.  The operation of electromagnetic flowmeters is
            based on the Faraday law of electromagnetic induction.   If flow of
            a conductive fluid in a pipe is normal  to a magnetic field, an
            electromotive force is induced across the fluid in a direction
            normal to both the magnetic field and the flow.  The induced
            voltage is measured by placing insulated electrodes across the
            pipe so that a line connecting them is perpendicular to the
            magnetic field.  See figure 1.
                      Figure 1.  Flowmeter schematic.


                 For an axisymmetric pipe flow and a uniform magnetic field
            the output voltage, E,  is given by
                      E = k B L V
[1]

-------
           where L is the distance between electrodes,  V is the area
           average velocity of the flow,  B is the magnetic flux density
           and k is a coefficient that is constant for  a given meter.
           The volumetric flowrate, Q, is then

                     Q =*D2 V/4                                        [2]

           where D is the diameter of the flow tube and is often, but
           not always, the same as L.

                The fluid being measured must be conductive.   The
           threshold conductivity is design dependent but is generally
           several micro-mhos per centimeter.

                The magnetic field is usually generated by the current alter-
           nating in the coils at line frequency.  However, in some meters
           the  line current is rectified and  imposed on the coils as a
           "pulsed direct current" alternating between a positive value and
           zero at a  frequency of  (typically) several  hertz.   The claimed
           advantage  of  these meters  is  that  each  pulse of current  lasts
           long enough for  an induced voltage to  be obtained  free of tran-
           sients and quadrature signals while the zero-current period is
           long enough for  the noise  signal  to be obtained and subtracted
           from the measured signal,  in  effect providing an internal zero
           correction with  concomitant reduction  of drift and  improved
           accuracy.

     3.1.2  Probe-type meters.  Although  this  report deals mainly with the
           tube-type  flowtneters  described  above,  the availability of other
           types of  electromagnetic  flowmeters is  noted here.  In these
           meters, an electromagnetic  sensor  measured  a "local" velocity to
           which the average  velocity in the conduit can  be related, prefer-
           ably by calibration.   The  following examples do not necessarily
           comprise  an  inclusive list.   They include:   cylindrical  velocity
            probes inserted into  the flow through  the conduit  wall or  flow
            surface;  "pitot" probes consisting of a short  submersible  flow
           tube intended for  use in large open or closed  conduits;  stream-
            lined sensors mounted at the  pipe invert near  a  manhole, for
            example.   These kinds of meters are  adaptable  to conduits  other
            than circular.  In the case of free surface flows  the depth must
            be monitored simultaneously.   The types of  fluids  in  which  they
            can be used depends in part on the intrusiveness of the  probe.

3.2  Advantages and Disadvantages of Electromagnetic Flowmeters

     3.2.1   Flow-tube type meters are non-intrusive.   This property  makes
            them particularly attractive  for use with sewage and sludge
            and also causes the head loss to be no greater than that for
            an  equivalent length of plain pipe of the same diameter.

-------
3.2.2  The output of tube-type meters can,  under certain conditions,  be
       made independent of pipe-roughness or Reynolds-number effects
       for axisymmetric flow;  but the electromagnetic flowmeter does
       not necessarily measure a true average velocity for all  velocity
       distributions.   See sections 5.1  and 6.2.  However, the  output
       is independent of fluid density and  is essentially linear.

3.2.3  In probe-type meters,  the sensing element can be readily withdrawn
       for monitoring and cleaning.  However, it may be sensitive  to
       changes  in velocity distribution.

3.2.4  Non-conductive liquids  and most gases cannot be metered  electro-
       magneti cally.

3.2.5  The instrument can be  subject to extraneous electrical effects
       that are not always obvious.

3.2.6  Servicing and trouble-shooting may require substantial equipment
       and skilled technicians.

-------
                       4.  GUIDELINES FOR SPECIFICATIONS

4.1  Flow Tube

     4.1.1  Geometry.

            4.1.1.1  The flow tube generally should have the same nominal
                     diameter as the adjoining pipe.   An exception can be
                     made when a smaller diameter is  necessary to maintain
                     a higher velocity through the meter (section 5.4.1).
                     Specifications for this exception are given in section
                     5.1.2.   There are no restrictions on the length of the
                     flow tube provided that a proper magnetic field can be
                     developed; generally the length  is greater than the
                     diameter.

            4.1.1.2  The manufacturer's literature should provide, for
                     design purposes, the actual internal diameter corres-
                     ponding to each nominal meter size.  If this actual
                     diameter is smaller than the upstream pipe diameter,
                     and if the flowing material is abrasive, a protective
                     orifice With diameter slightly smaller than that of the
                     liner can be installed at the upstream flange to protect
                     the leading edge of the liner.  To insure that perfor
                     mance is not affected, calibration should be done with
                     the protective orifice in place.

            4.1.1.3  Flow tubes are usually flanged at the ends.  Because  the
                     inner lining is usually carried  around to the flange  face
                     to provide insulation, gasketing and torquing of the
                     flange bolts become critically important; the manufacturer
                     should provide detailed instructions and all necessary parts
                     to avoid liner damage on installation.   For flow tubes that
                     are not flanged, instructions should similarly be provided
                     as necessary.

            4.1.1.4  Direction of flow must be indicated on the outside of the
                     flow tube, unless the meter is completely bi-directional
                     without adjustment.

     4.1.2  Materials.

            4.1.2.1  The flow tube must be made of non-magnetic material and
                     the inner liner in contact with  the flow must be non-
                     conductive as well.   The materials must be unaffected
                     by line pressure.

-------
       4.1.2.2  The liner material  also must be suitable for the flowing
                liquid in terms  of  resistance to corrosion, abrasion,
                etc.   In  some  cases,  glass or Teflon lining may be
                advantageous for sludge flow.  Liners used for digested
                sludge should  be capable of handling temperatures to
                70°C.   All  liners should be able to resist low tempera-
                tures  typical  of the  location.
4.1.3  Electrodes.
       4.1.3.1   Electrodes  are  located diametrically opposite each
                other,  and  are  either flush with the inner tube surface
                or  protrude slightly into the flow.  See figure 2.
                They must be  insulated from the remainder of the flow
                tube.   The  protruding configuration is used to hydro-
                dynamically increase the velocity around the electrode
                and enhance its self-cleansing capability.  A protruding
                electrode should be rounded at the edge so that a
                curvilinear protuberance is presented to the flow.
                Electrode surfaces should never be recessed, since the
                resulting cavity may collect debris.
                   Eledrode

                                 (b)
                    Figure 2.   Electrodes

-------
            4.1.3.2  Electrodes  must be made of corrosion-resistant material.

            4.1.3.3  It is preferable but not mandatory that the electrodes
                     be removable without taking the meter out of the line.

            4.1.3.4  Small (in area) electrodes are recommended.  In
                     principle,  the sensitivity of the flowmeter to velocity
                     distribution can be eliminated by the use of large
                     electrodes  rather than the conventional or "point"
                     electrodes.   However, the problems introduced in
                     keeping large electrodes clean are likely to outweigh
                     the potential advantages and they are not recommended
                     for wastewater treatment plants.

            4.1.3.5  Internal  electrode cleaning capability should be built
                     in, particularly when frequent electrode inspection is
                     not feasible.  This internal cleaning can be accomplished
                     by mechanical scrapers, ultrasom'cally induced vibrations,
                     heating,  or by other means.  It can be either continuous,
                     intermittent or manually actuated.  Continuous operation
                     is recommended for ultrasonic cleaners, since experience
                     suggests  that grease deposits, once formed, are difficult
                     to remove by this method.

     4.1.4  Grounding.   All  necessary grounding straps must be furnished with
            the meter.   Complete instructions for earth grounding and methods
            for ascertaining the quality of the ground must be furnished.  See
            also section 5.5.4.

4.2  The Secondary

     4.2.1  The secondary system can be mounted integrally with the flow tube
            or separately from it.  If it is mounted separately from the primary,
            the manufacturer must furnish specifications for all  electrical
            connectors  along with complete instructions, distance restrictions,
            etc.

     4.2.2  The flow converter must include a display  either in flowrate units
            or in units that can be readily converted  to flowrate.   It is  prefer-
            able that the readout have switch-selectable flow ranges;  however,
            the zero offset should be checked for each range.   A flowrate
            display should be  mounted near the flowmeter;  this is in addition
            to any flowrate readout or flow totalizer  in the central  control
            room.

     4.2.3  The secondary should be capable of being sealed against excessive
            humidity.   In some instances,  it will  be necessary to specify
            that all  meter elements be able to withstand temporary immersion
            for periods up to  48 hours at depths determined by the depth of
            pits or vaults that  may be accidentally flooded.

-------
     4.2.4  Electrical  connections and components must satisfy all  code
            requirements for the particular area classification.

     4.2.5  The secondary system should be able to operate within an ambient
            temperature range of -30°C to 60°C.

4.3  Calibration

     4.3.1  The flowmeter should be hydraulically calibrated by the manufacturer
            prior to delivery to the customer.

            4.3.1.1   The conditions under which the factory calibration was
                     performed, including upstream geometry, should be made
                     known to the user.

            4.3.1.2  If a factory hydraulic calibration was not made, for
                     whatever reason, the alternate means by which  the meter
                     rating was determined shall be fully explained to the
                     user.

     4.3.2  The manufacturer should furnish, or make available as an option,
            equipment for calibrating the flowmeter from the electrodes to
            the readout, or shall alternatively furnish instructions for
            performing  such a calibration of the secondary system.

     4.3.3  For probe-type flowmeters, the calibration data should  include
            information on the effects, if any, of the radial probe location
            with regard to changes in pipe roughness or Reynolds  number.

4.4  Accuracy

     4.4.1  The accuracy of an electromagnetic  flowmeter should be  within  one
            percent of  full scale, but should not exceed three percent of
            flow.  This means that if the flowmeter is to be used at flowrates
            Tower than  about one-third of full  scale,  a tighter accuracy
            specification will have to be put on the meter.

     4.4.2  The repeatability shall be within one-half percent of full  scale.
            For process control purposes, repeatability is  often  more important
            than accuracy.  On the other hand,  for measurements involving
            discharge-permit requirements or allocation of treatment costs  or
            capacity among jurisdictions or industries, accuracy  and repeat-
            ability are equally important.

     4.4.3  The output  should be independent of variations  of up  to 10 percent
            in line  voltage and frequency,  and  of temperature variations within
            the ranges  cited in sections 4.1.2.3 and 4.2.5.

     4.4.4  If the manufacturer claims that electrodes and/or secondary elements
            are completely interchangeable, the accuracy limits for the system
            with interchanged parts should  be furnished to  the user.

-------
                         5.   INSTALLATION REQUIREMENTS

5.1  Approach Conditions

     5.1.1  At least five diameters of straight pipe, measured from the plane
            of the electrodes,  should precede and follow the tube.   However,
            this approach length does not insure freedom from upstream effects,
            since errors described in section 6 can be introduced by severe
            upstream disturbances.  If approach lengths shorter than 5 dia-
            meters are recommended for a specific commercial meter, data
            and/or a description of the experiments or analysis on  which this
            recommendation is based should be furnished to the user.

     5.1.2  An exception to  the requirement of section 5.1.1 can be made in
            the case of a tapered axisymmetric reducer which is installed imme-
            diately upstream of a smaller flow tube in order to provide a self-
            cleansing velocity  within the flow tube.   The included  angle of
            such a taper should not exceed 30 degrees, both upstream and down-
            stream of the flowmeter.

5.2  Bypass

     5.2.1  It 1s strongly recommended that piping and isolation valves be
            installed in such a way that flow can be  bypassed around the meter.
            This arrangement not only permits removal of the meter  when ser-
            vicing is necessary, but also allows for  convenient stopping of
            flow through the meter for zero checks.   It is important that
            there be no leakage flow through the valves.

     5.2.2  It is also recommended that a cleanout tee be installed near the
            flow tube.  See  figure 3 for one suggested layout.   This accessi-
            bility is particularly important for meters that do not have
            electrode-cleaning  capability or are subject to deposition or
            coating.

5.3  Orientation and Location

     5.3.1  It is preferable that the flowmeter be installed in a vertical
            length of pipe.   This orientation tends to equalize wear on the
            lining when an abrasive liquid is flowing and also prevents gases
            from collecting  at  the pipe crown.

     5.3.2  If the flowmeter cannot be placed in a vertical  pipe and must be
            inserted in a horizontal  or inclined line, the following steps  are
            recommended.
                                    10

-------
                                     ISOLATING

                                      VALUE.
                     BY-PASS VALUE
                                               ISOLATING VALUE
                                           FLOW
                      NOTE: RECOMMENDED 5 D UP AND DOWN STREAM
                      Figure 3.  Typical  installation.

            5.3.2.1  Orient the meter so  that  the electrodes  are  on  a
                     horizontal line.

            5.3.2.2  If the pipe is horizontal,  install  bleed valves at the
                     crown near the flowmeter  so that  the  line  can be
                     checked for accumulated gas.

            5.3.2.3  In any event the flowmeter  should always be  located so
                     that it will remain  full, both  during flow and  when the
                     flow is stopped.  See also  section  6.5.2.

5.4  Limiting Velocities

     5.4.1  To assure scouring action, design  for minimum  velocities of at
            least 1.5 m/s (5 ft per second) for  primary  sludge  and at  least
            0.9 m/s (3 ft per second) for other  sludges  and raw influent
            sewage.

     5.4.2  To avoid excessive abrasion,  velocities  should never  exceed 8
            m/s (25 ft per second).

5.5  Other Considerations

     5.5.1  Place the flowmeters on the high pressure  side of pumps  and
            control valves; this will minimize the effect  of,gas  bubbles.
                                       11

-------
5.5.2  Avoid proximity to heavy induction machinery.

5.5.3  Refer to section 4.1.1.3 for mechanical  installation procedures.

5.5.4  Proper grounding is essential (see section 4.1.4);  improper ground-
       ing is one of the most frequent causes of failure of electro-
       magnetic flowmeters in wastewater treatment plants.   Grounding
       requirements may vary, depending for example on whether the pipe
       is metallic or nonmetallic,  or on the meter material.   An internal
       grounding ring (which also provides abrasion protection)  may be
       required.  In all cases the  manufacturer's instructions must be
       followed closely.
                                12

-------
                             6.  ERROR SOURCES

6.1  General.  The errors considered in this section are over  and  above  the
     basic accuracy inherent in the instrument under optimum conditions  of
     flow and use; this basic accuracy (as defined  in 2.1)  shall not  exceed
     the limits given in section 4.4.  It can be  less than  these limits  if
     so indicated by in-place calibration or by complete data  furnished  by
     the manufacturer.

6.2  Velocity Profile

     6.2.1  Background.

            6.2.1.1  It has been shown theoretically that  it is  impossible
                     to design an electromagnetic flowmeter with "point"
                     electrodes that will be completely insensitive to all
                     types of velocity profiles (1).  Compromises  can be
                     attained as given in the following.

            6.2.1.2  It has been shown theoretically that meters designed
                     with a uniform magnetic flux are insensitive  to  the
                     velocity profile provided it remains axisymmetric (2).
                     This means that changes in viscosity or in  (uniformly
                     distributed) pipe roughness will not affect the
                     flowmeter response.  But in order to attain
                     axisymmetry, long straight upstream approach  piping is
                     needed.

            6.2.1.3  Commercial electromagnetic flowmeters  are designed with
                     non-uniform magnetic flux density in an attempt  to even
                     out the weight function and render the meters less
                     sensitive to axisymmetric velocity distributions.  This
                     may result in losing the inherent insensitivity  to
                     axisymmetric changes described in section 6.2.1.2;
                     however, this effect will generally be small  -- well
                     under 1 percent (1) of full scale.

            6.2.1.4  Only limited data based on systematic  experimental
                     investigation of upstream effects are  available  in the
                     literature, and even these show some disagreement.
                     Sensitivity to upstream conditions depends  in part upon
                     meter design, and experimental results are  not
                     necessarily transferable to other types of meters.
                     Nevertheless, the recommendations in section  6.2.2 are
                     proposed as practical guidelines based on published
                     information.
                                      13

-------
     6.2.2  Effects of upstream pipe fittings.

            6.2.2.1  For the purpose of this section, a straight  approach
                     length of 10 diameters will be considered  to  restrict
                     errors due to velocity distribution to less  than 1
                     percent and will be treated as an essentially
                     error-free condition (3).

            6.2.2.2  Axisymraetric fittings, e.g., expanders, reducers,  etc.,
                     located at least 5 diameters upstream of the  electrode
                     plane probably will not measurably affect  the perfor-
                     mance, provided that any asymmetric fittings  in the
                     line are at least 5 diameters farther upstream.
                     Fittings that introduce asymmetric disturbances, e.g.,
                     elbows, gate valves, etc., and are located 5  diameters
                     (and up to 10 diameters) upstream can be estimated to
                     add an additional 1 to 3 percent uncertainty  to the
                     flow (4).  Fitting, even closer to the electrode plane
                     can cause substantial errors.  For example,  a half-
                     closed gate valve located 1 diameter upstream can  cause
                     errors of 8 percent and higher (4).  It is not possible
                     to further quantify errors from these sources using
                     published information.  The size of the error may  also
                     depend upon the orientation of the eccentricity
                     relative to the line connecting the electrodes.

            6.2.2.3  The effect of swirl has not been established;  an
                     uncertainty of 1 percent should be allowed (5).

            6.2.2.4  There is little information on effects of  downstream
                     fittings; a straight length of 3 diameters is  suggested
                     for asymmetric fittings.
6.3  Deposits
     6.3.1  Deposits of non-conducting material on the electrodes will tend
            to Insulate the electrodes and cause erroneous readings.  Such
            deposits can be minimized by electrode shape (section 4.1.3.1),
            automatic cleaning capability (section 4.1.3.5), or maintenance
            of high flow velocities.

     6.3.2  Deposits of non-conducting material on the tube liner (but not
            on the electrodes) will not affect the result unless the deposit
            becomes thick enough to measurably change the tube diameter and
            affect the output obtained from equation [2].  Such accumula-
            tions are discouraged by maintenance of high flow velocities.

     6.3.3  Deposits of non-conducting material on the (originally
            conducting) pipe immediately upstream or downstream of the
            flowmeter will  cause the liquid to lose its grounding.  Where
            this problem is anticipated, the grounding should be done with
            grounding spikes or rings (or their equivalent) protruding into
            the flow.
                                      14

-------
     6.3.4  The effect of conductive deposits within the flow tube depends
            in part on the ratio of the conductivity of the deposit to that
            of the liquid.  Deposits of the same conductivity as the liquid
            may introduce no error except for that associated with change of
            diameter.   However, conductive deposits can cause shorts to
            develop between the electrodes, or between the electrodes and
            ground, and they should be discouraged by the use of high
            velocities, elevated temperatures, or frequent cleaning.

6.4  Air or Other Gases

     6.4.1  Air or gas bubbles diffused through the flow will not introduce
            an error in the volumetric flowrate measurement.   However,
            mass flowrates deduced from these measurements will  be in error.
            See section 5.5.1.  If gas collects so that the pipe is not
            full (as distinguished from bubble dispersion), an error will be
            introduced even if the electrodes remain submerged.   See section
            5.3.

     6.4.2  Small gas  bubbles effervescing from the liquid and adhering to
            the electrodes will introduce errors by partially insulating
            the electrodes.  Maintenance of recommended velocities (section
            5.4.1) should preclude this occurrence.

6.5  Zero Drift

     6.5.1  Zero drift is a commonly occurring error source in electro-
            magnetic flowmeters, and it can be monitored only by frequent
            checking with the flow stopped.  Flowmeters that  have built-in
            zero correction as described in section 3.1.1 may require only
            infrequent monitoring.   See section 7.1.2.7.

     6.5.2  There is limited evidence that alternate immersion and
            drying of  the electrodes encourages the formation of thin
            coatings that contribute to zero drift (5).  However, keeping
            the flowmeter tube full at all times, even during extended
            periods of no flow, is suggested only for clean flows in
            non-corroding lines.  In view of the proclivity of sewage and
            sludge to  form grease deposits, these flows should be stopped
            under meter-full conditions only long enough for  zero checks.
            A clean water flush is required following longer  term shut down.
                                       15

-------
                    7.  CALIBRATIONS AND PERFORMANCE CHECKS

71.1   Hydraulic Calibration of the Measuring System

     7.1.1  General.

            7.1.1.1  Even though a flowmeter is delivered pre-calibrated
                     hydraulically by the manufacturer in accordance with
                     section 4.3.1, an in-place hydraulic calibration
                     provides the most assured method of evaluating its
                     performance.
                     - Hydraulic in-place calibration is recommended when
                       the system is first installed, at which time it
                       provides not only an initial  calibration but also an
                       acceptance test.
                     - Hydraulic calibrations are recommended also at times
                       subsequent to the initial  calibration, if electrodes
                       or secondary elements are  changed, or if monitoring
                       indicates an unexplained change in the measured flow-
                       rate.
                     - General  procedures for hydraulic calibration are
                       described in section 7.1.2.   Methods of obtaining
                       reference flow measurements  for hydraulic calibrations
                       are given in section 7.3.
                     - Exceptions to the requirement for hydraulic 1n-place
                       calibration are given in section 7.2.

            7.1.1.2  It is emphasized that even an  in-place hydraulic
                     calibration reflects the performance of an electro-
                     magnetic flowmeter only so long as the electrodes
                     and the tube interior are not  adversely affected and
                     remain effectively in the same  condition as during
                     calibration.  Therefore, continued monitoring and
                     follow-up  calibrations are important,  especially in
                     sewage and sludge flows.

     7.1.2  General procedure for hydraulic calibration.

            7.1.2.1  Be sure  that the electromagnetic flowmeter has been
                     set up and that preliminary  checks have been made
                     completely in accordance with manufacturer's instruc-
                     tions.

            7.1.2.2  With  the line full  but with  the discharge valves closed
                     (pump,  if  not positive displacement,  should be running)
                     check the  zero-flow reading  of  the instrument.   If there

                                      16

-------
         is a zero offset, correct it according to manufacturer's
         instructions.

7.1.2.3  Set up a flowrate through the meter that is within the
         anticipated working range of flows.  Allow enough time
         for it to become steady, and measure the reference or
         "true" flowrate, QR, by one of the methods in section 7.Z.3.

7.1.2.4  During the time that the measurement of QR is being made,
         the flowrate as measured by the electromagnetic flowmeter,
         Q, should be recorded.   If the measurement of QR is made
         by a method that requires a substantial length of time,
         e.g., the volumetric method, readings of Q should be
         made at intervals during the test period and an average
         value determined.

7.1.2.5  Compute the percentage  difference in the two flowrates
         from 100 (Q-QR)/Qn and  enter this value on an error vs.
         flowrate curve as shown in figure 4.

7.1.2.6  The procedure  of sections 7.1.2.2 through 7.1.2.5
         should be done for a minimum of three flowrates (low,
         medium and high) within the anticipated working range.
         It is preferable that numerous runs be made (above the
         minimum three) to provide an indication of repeatability
         as shown in figure4 .   (As a practical matter, however,
         it is recognized that some of the calibration flow
         measurements of section 7.2 are very time and resource
         consuming and  that repeated runs may not be feasible.)
         Draw curve A averaged through the points.

7.1.2.7  A zero check (section 7.1.2.2) is recommended between
         each run in order to check the short-term zero drift.
         - If there is  substantial zero drift, it should be
           ascertained  whether it is due to local conditions,
           such as changes in nearby inductive machinery, which
           should not necessarily be charged against the flowmeter.
         - If the zero  drift appears to be characteristic of
           the instrument itself, it should be monitored before
           each test point but no adjustment should be made.
         - Even if a flowmeter has automatic internal  zero
           correction,  a no-flow reading is desirable  as an
           initial check on the  correction.

7.1.2.8  Estimate the percentage error in measuring QR and enter
         it as curve B  in figure 4.
         - This error estimate for QR is especially important
           in acceptance tests,  since the flowmeter under test
           cannot be held to an  accountability stricter than the
           user's ability to check it.
         - The estimated error will clearly depend upon which
           method of section 7.3 is used and the manner in which
                          17

-------
QJ
O)
_O
U.
       5


       4


       3


       2
      0
     -3
                                       t  Q
                                                                  8
       Figure 4.  Flowmeter error determination.
                            18

-------
                     it is executed.  It may or may not vary with  flowrate.
                     Some of the methods, e.g., several of the  transfer-
                     standard devices, have documentation that  provides
                     guidelines for estimating errors  (7,8).  In any  event,
                     estimated errors that are agreeable to the involved
                     parties must be assigned to QR.

            7.1.2.9  If curve A is outside of curve B  by more than  the
                     amounts allowed in section 4.4.1  (or alternate specifi-
                     cations), the performance may be  unsatisfactory  and
                     these options should be considered.
                     - If the test results in figure 4 suggest  a constant
                       percentage error, it may be that a span  adjustment  of
                       the secondary is required.  The manufacturer's manual
                       should provide instructions by  which the user  can
                       make this adjustment.  The calibration process of
                       section 7.1.2 should be repeated after such  an
                       adjustment.
                     - If the flowmeter cannot be brought within specifica-
                       tions with adjustments made by  the user,  it  must be
                       repaired or rejected.  However, if the data
                       differences can be ascribed to  faulty approach
                       conditions and/or are otherwise consistent and
                       repeatable, the calibration results can  be used to
                       develop a new, in-place rating  for the flowmeter.

           7.1.2.10  The error comparison described in sections 7.1.2.8 and
                     7.1.2.9 should be regarded as a suggested  procedure
                     only.  Other comparison procedures that are agreeable
                     to the involved parties are acceptable; the important
                     point is that the reliability of  the reference measure-
                     ment must be taken into account in some way.

7.2.  Calibration Other Than Hydraulic

      7.2.1  It is recognized that there are situations where in-place
             hydraulic calibrations are not feasible and the user must
             resort to other methods.

            7.2.1.1  One alternative is to calibrate the secondary  only,
                     using an electronic "flow simulator" offered as an
                     option by the manufacturer or following instructions
                     provided by the manufacturer to accomplish  the same
                     result.  (See section 4.3.2).  This alternative  should
                     be used in a place of a complete  calibration only when
                     there is reasonable certainty as  to the condition and
                     performance of the primary.  This requires (assuming
                     the flux strength can be verified) that the electrode
                     and flow tube surfaces be closely monitored and that
                     the adjacent piping is suitable.


                                     19

-------
           7.2.1.2  If there are upstream pipe fittings, valves, etc.,
                    closer than 5 diameters to the electrodes, and particu-
                    larly if these fittings produce asymmetric flows, the
                    calibration should include the entire measuring system,
                    i.e., primary as well as secondary.

    7.2.2  Situations such as those described in section 7.2.1.2 can be
           approached also by "dry calibration" techniques, which involve
           direct measurement of the magnetic field strength in combination
           with weight functions and velocity distributions to arrive at a
           flowrate.   Details of this method are beyond  the scope of this
           report.  See, e.g., (6).  The velocity-distribution data needed
           for this method normally would be difficult to obtain, particu-
           larly in a sludge.  However, it may be feasible to estimate a
           "worst case" velocity distribution based on published information
           so that error limits can be estimated.

7.3  Flow Measurement Methods for Hydraulic Calibrations

     7.3.1  General.

            7.3.1.1  The purpose of section 7.3 is to provide a general
                     overview of methods available for obtaining the
                     reference flowrate, QR» needed in section 7.1.   The
                     methods most likely to be applicable here are:
                     - Volumetric
                     - Comparison with a reference flowmeter
                     - Dilution
                     - Salt velocity
                     - Velocity traverse

            7.3.1.2  The methods described vary considerably in accuracy and
                     difficulty.  Advantages, disadvantages, and areas of
                     application of each method are cited.   The method
                     selected by the user will depend upon the purpose of
                     the calibration, the type of liquid flowing, and the
                     resources available to conduct the  tests.

            7.3.1.3  For highest accuracy, the in-place  hydraulic calibration
                     should be conducted with the actual  process liquid.

            7.3.1.4  If the calibration flowrate measurements are made at a
                     location in the plant that is not directly in the line
                     containing the electromagnetic flowmeter, equivalence
                     of the flowrate at the measurement  location to  that
                     through the flowmeter must be assured.

     7.3.2  Volumetric calibration.

            7.3.2.1  The feasibility of volumetric calibration of course depends
                     upon the availability of suitable tank space and connecting
                     conduits.   Of the methods described here, it is probably

                                     20

-------
                the one most suitable for difficult sludges.   The
                potential  accuracy is high, provided that:
                - The tank is regular in configuration so that its
                  lateral  dimensions can be measured within acceptable
                  limits of accuracy.
                - The tank is large enough to permit a test run of
                  sufficient length for the effect of timing errors at
                  the start and finish to be kept within acceptable limits.
                - The change in liquid level during the run is large enough
                  so that the starting and finishing depths (probably
                  obtained by the "on-the-run" method) can be measured
                  within acceptable relative error limits.
                - The flowrate remains relatively constant during the
                  run.   (See section 7.1.2.4.)
                - It is noted that gravimetric tests can be substituted
                  for volumetric where suitable tanks and scales are
                  available.

       7.3.2.2  Estimate the uncertainty of the resulting QR from a
                combination of the estimated errors of measurement of
                the lateral area, the depth change, the elapsed time.
                This uncertainty should be used in figure 4 as
                indicated in section 7.1.2.

7.3.3  Comparison with a reference meter (transfer standard).

       7.3.3.1  In this context a reference meter is a flowrate measuring
                device whose performance can be referenced to published
                standards or to recommended practices that are acceptable
                to the involved parties.  Examples include:
                - Standard venturi tubes and venturi nozzles  (7,8)
                - Orifice plates (7,8)
                - Parshall flumes (9)
                - Thin plate weirs (10)

       7.3.3.2  Flowmeters used as reference devices must meet all
                requirements of the accepted standard practices in
                fabrication, installation and use, so that their flow
                coefficients and uncertainties can be used for determining
                Qn.  In practical treatment-plant situations, it may be
                difficult to find or install a reference flowmeter  meeting
                all of these requirements.   For example, the  stringent
                upstream approach conditions may not be met.   In that case,
                the flowmeter cannot be used as a transfer standard
                unless  rationally based and defensible modification to
                the flow coefficients and uncertainties can be agreed
                to by the  involved parties.

       7.3.3.3  When a  differential-pressure type of flowmeter is used
                as the  reference device, measure the pressure difference
                with a  U-tube manometer.  If a commercial  secondary
                device  is  used in place of a manometer, it must have had

                                 21

-------
                a recent calibration and  complete  information  on  its
                performance must be available;  its error  should be
                included in the QR uncertainty.

       7.3.3.4  When a standard weir or Parshall flume  is  used as the
                transfer flowmeter, measure  its  head with  a  point gage
                (carefully zeroed to the  crest)  or equivalent  device
                and determine QR from standard  equations.  If  a
                commercial secondary device  is  used, see  section
                7.3.3.3.

       7.3.3.5  It may be acceptable to use  as  a reference device a
                flowmeter for which there are no published standards,
                e.g., acoustic flowmeters, segmental orifices,
                provided:  the device has been  recently calibrated  and
                its current accuracy and  repeatability  can be
                quantified to the satisfaction  of  involved parties; and
                the device is used under  effectively the  same
                conditions for which it was  calibrated.

7.3.4  Dilution method.

       7.3.4.1  In the dilution method the flowrate is  deduced from the
                dilution of measurable properties  (e.g.,  color, conduc-
                tivity, or fluorescence)  or  tracer chemicals added  to a
                turbulent flow in known amounts.  The calibration can
                be done by either the constant-rate injection  method,
                or the slug injection method.  The constant-rate  method
                is recommended here because  it  appears more practical
                for In-plant use.  See reference (11) for  details.

       7.3.4.2  In the constant rate Injection  method, a  tracer solu-
                tion of accurately known  concentration 1s  Injected
                upstream at a rate which  1s  constant and  accurately
                measurable.  At a downstream distance long enough for
                complete mixing, the flow is sampled and  the concentra-
                tion "plateau" is attained.  The flowrate, Q,  Is  then
                determined from

                         Q = q(c] - c2)/(c2  - c0)          [3]

                where: q is the rate at which the sample of concentra-
                tion C] is Injected; c2 is the measured "plateau"
                concentration downstream; and c0 (which may be close
                to zero) is the background concentration of the tracer
                chemical existing in the  flow.

       7.3.4.3  This method requires accurate measurement  of q and  of
                all concentrations; skilled  personnel and  specialized
                equipment are needed.  However,  under optimum  condi-
                tions the potential accuracy is  high.  See reference
                (11) for methods of estimating errors.
                                  22

-------
       7.3.4.4  The tracer property must be conservative,  since losses
                by absorption to solids in the flow will  result in  an
                apparent reduction in Cp.   The fluorescent dye
                Rhodamine WT has been used successfully in sewage
                without losses;  its behavior in sludge is  not known.

7.3.5  Salt-velocity method.

       7.3.5.1  In the salt-velocity method, brine is  injected suddenly
                at an upstream station in such a way that  it becomes
                well-distributed across the section very  rapidly.  The
                time of passage  of the salt pulse between  two downstream
                stations is measured by means of electrodes which detect
                the increased conductivity associated  with the passage
                of the brine. The flowrate then can be determined
                provided the volume of the conduit between the electrodes
                is accurately known.  This method has  a potential for
                1  percent accuracy under optimum conditions.   The
                accuracy actually obtained depends upon the tranverse
                mixing and coherence of the injected brine slug,  upon
                the accuracy of  determination of the centers of gravity
                of the tracer conductivity and the time separating  them,
                as well as upon  the accuracy of the aforementioned
                volume determination.

       7.3.5.2  Published standards for the salt-velocity  method  are
                written for circular pipes flowing full  (8,12),  and these
                or similar references must be consulted for details of
                the method.   A sufficient length of (preferably  straight)
                pipe upstream of the first electrode is necessary to insure
                complete lateral mixing of the salt slug when it  reaches
                the electrode.   This length can be short as four  diameters
                when the injection is done internally  in the  standard
                manner (8,12).   However, a substantially longer approach
                distance is needed if injections are made  from the periphery
                of the pipe.  The distance between the two sets of electrodes
                must be at least four diameters.

       7.3.5.3  The liquid being measured  must have a  significantly
                smaller electrical  conductivity than the brine.

       7.3.5.4  The brine injection must be sudden,  with an injection
                interval of the  order of 1  second and  no leakage
                thereafter.

       7.3.5.5  The electrodes must provide equal  increments  of
                conductivity for equal  segments of cross-sectional
                area.   This  requires the electrodes  to be  intrusive,
                so that the  method  cannot  be used in raw sewage or
                sludge.   However,  it is  suitable for treated  effluent.
                                 23

-------
       7.3.5.6  In principle, this method can  be  adapted  to  other
                shapes of conduits or channels provided that the
                approach length and the electrode spacing and confi-
                guration are modified to compensate  for the  shape
                change in a manner that is  hydrodynamically  sound  and
                agreeable to the involved parties.

7.3.6  Velocity-area method.

       7.3.6.1  The velocity-area method is applied  to a  flow cross
                section by measuring a number  of  velocities  over the
                section, each representative of the  average  velocity
                within an incremental area, and summing the  resulting
                velocity-area products.  The method  can be applied to
                both open and closed conduit flows,  but it is much more
                convenient in accessible open  channels.

       7.3.6.2  This method 1s likely tp be suitable mainly  for elec-
                tromagnetic flowmeters of relatively large capacity
                discharging a fairly clean  liquid.

       7.3.6.3  The velocity can be measured by point-velocity measur-
                ing instruments such as current meters (rotating or
                electromagnetic), pltot tubes, etc., or by acoustic
                velocity meters that measure an average velocity
                component along a line path.

                - The point velocity instruments  are intrusive and
                  generally could not be used  effectively 1n  raw sewage
                  or sludges.  However, there may  be an opportunity to
                  use a rotating-element or electromagnetic  current
                  meter 1n an open channel  discharging treated
                  effluent.  In that event, see section 7.3.6.5.
                - Pltot tubes are usually restricted to full   conduit
                  flows, where velocities are more likely to  be high
                  enough to generate a readily measurable dynamic
                  pressure.

       7.3.6.4  The accuracy of this method depends upon whether the
                sampling points are distributed so as to yield an
                average velocity and whether each  velocity is  accurate-
                ly measured.  The latter depends  in part on  the
                accuracy of the velocity-measuring instrument  itself
                and upon the duration of the sampling time at  each
                point.   The sampling requirements  tend to make it  a
                lengthy measurement,  so 1t can be  used only where
                sufficiently long periods of essentially  steady flow
                are available.

       7.3.6.5  In cases where a point-velocity instrument is  used in
                an open channel,  the following conditions must be
                observed.
                -  The average velocity in the section preferably should
                  exceed 0.30 tn/s (1  ft per second).


                                 24

-------
                     - Only velocity-measuring instruments that have been
                       recently calibrated and whose present accuracy and
                       uncertainty can be estimated to the satisfaction of
                       involved parties should be used.
                     - Consult reference (12) for distribution of velocity
                       sampling points in the cross section, and reference (13)
                       for error estimates.

7.4  Flow Measurements for Monitoring

     7.4.1  General.

            7.4.1.1   Measurements for monitoring are defined in this context
                     as those that are made subsequent to an initial calibration
                     for the purpose of detecting anomolous behavior of the
                     electromagnetic flowmeter.  These measurements in
                     themselves cannot furnish a calibration coefficient as
                     can those listed in section 7.3.  Nevertheless, the
                     measurements must be carefully made so that their
                     precision can be depended upon.

            7.4.1.2   Frequent monitoring is especially important for users
                     of electromagnetic flowmeters, since the system is
                     electrical as well as generally obscured from view and
                     consequently the quality of its performance is not as
                     visible and readily checked as, say, an open channel
                     flume with a float gage.

            7.4.1.3   Measurements for monitoring frequently involve
                     measuring a pressure difference at a location where
                     the flowrate has a unique and repeatable relationship
                     to that difference so long as the geometry remains
                     unchanged.  This requirement for constant geometry means
                     that, in the case of valves, the settings for disc or
                     gate position must be reproducible.   It also means that
                     there cannot be internal deposits or coatings that
                     significantly affect the measured pipe diameter.
                     Locations for monitoring can include, among others:
                     - Pipe elbows, usually 90 degrees (see also section 7.5.2)
                     - Butterfly or gate valves (when partly closed)
                     - Pumps
                     - Pipe reducers

            7.4.1.4   The capability for monitoring should be set up prior to
                     calibration, so that a  relation between the pressure
                     difference at the monitor and the reading of the  newly
                     installed electromagnetic flowmeter can be established
                     from the results of the hydraulic calibration (section
                     7.1.2) or by other means (section 7.2).

            7.4.1.5   It should be noted that this type of monitoring
                     procedure is valid only so long as the viscosity  of the

                                       25

-------
                liquid does not change radically; caution should be
                exercised with sludges.

7.4.2  Measuring the pressure difference.

       7.4.2.1  Pressure taps.

                - Pressure taps should be  flush with the inner surface
                  of the pipe and free of  burrs.  Hole diameter is not
                  critical because we are  looking for repeatability
                  rather than accuracy;  diameters of from 0.5 cm to
                  1.0 cm for small to large pipes are usually adequate.
                  The condition of the inner surface at the tap must
                  remain essentially the same and should be monitored
                  to that end.  For example, rust or grease ridges or
                  tubercles at or near the hole will affect the reading.
                - Pressure taps for elbows are installed as shown in
                  figure 5.  For the other locations cited in
                  section 7.4.1.3, the taps are placed upstream and
                  downstream of the device producing the pressure
                  difference.
                - For sewage and sludges it is recommended that only
                  one pressure tap be used at each upstream and
                  downstream point.  (Elbows use only one tap at each
                  position in any event.)   The use of several taps
                  evenly spaced around the circumference and connected
                  by a piezometer ring is  preferably restricted to
                  relatively clean liquids.
                - Single pressure taps in  horizontal or near-horizontal
                  lines should be located  in a horizontal  diametric
                  plane, or slightly above that position,  in order to
                  minimize entry of gases  and solids into the manometer
                  lines.

       7.4.2.2  Differential-pressure sensor.

                - The pressure difference  can be measured using
                  conventional U-tube manometry, for example, a water-air
                  manometer or, for large  differences, a water-mercury
                  manometer.  Manometer fluids should be selected so
                  that the manometer deflection is never less than
                  approximately 3 inches (7.6 cms).   The glass tubes
                  should be large enough to avoid meniscus shape errors.
                - The pressure difference  can be measured also with one
                  of various types of commercial differential pressure
                  cells.  Such sensors should be calibrated frequently
                  against a liquid-column  manometer.  This is best
                  accomplished by providing fittings in the connecting
                  tubing so that a manometer can be placed conveniently
                  in parallel with the sensor.
                - Always bleed manometer lines of gas before each use,
                  and check for zero deflection before and after each use.


                                  26

-------
                     - Make several  manometer readings (more,  if the columns  are
                       oscillating)  for each measurement and use the average.

            7.4.2.3  Connecting tubing.

                     - The recommendations  for the connecting  tubing between
                       the pressure  taps and the manometer or  sensor are
                       intended to discourage accumulations of gas or
                       sediment in the lines.
                     - To that end,  lengths of horizontal  tubing should be
                       avoided in favor of  moderate slopes.  Further,
                       elevation peaks (inverted U's)  in liquid lines
                       should have gas bleed valves,  and troughs should
                       have sediment drains or traps.
                     - The connecting tubing should be corrosion resistant
                       and its bore  should  be at least 3/8 inch (1 cm).  It
                       should be valved in  a way that permits  all  portions
                       of the lines  to be flushed when necessary.
                     - Valves and by-pass tubing should be provided so that
                       the manometer or sensor can be isolated from the pipe
                       flow for zero checks.
                     - Clean water should be kept in the lines by the purge
                       flow (see section 7.4.2.4).

            7.4.2.4  Purge flows.
                                                                             *
                     - Continuous purging of pressure taps is  required in
                       sewage and sludge flows.   This  purging  should be
                       done with tap water  or clean non-potable water.
                     - The pressure  loss in the  tubing between the purge-
                       water connection and the  tap should be  the  same in
                       both of the lines so that the pressure  differential
                       is essentially unaffected by the purge  flow.   This
                       can be accomplished  by making the two paths geometrically
                       similar and by keeping the purge flowrate the same  in  both
                       legs.  Install a variable area  flowmeter (or equivalent)
                       and a control valve  in each of the purge water lines for
                       flow adjustment.
                     - Purge flows should be high enough to keep dirty
                       liquid out of the connecting lines  under all
                       conditions; however  they  should remain  negligible
                       relative to the primary flow.

7.5  Approximate Flow Measurements

     7.5.1   It is useful  on occasion to have a relatively quick and
            inexpensive way of knowing whether a flowmeter is  even
            approximately correct, particularly  when an in-place calibration
            cannot be made.   Measurements in the range of +_ 10 percent
            uncertainty may be adequate for this purpose.   Some of the
            monitoring locations listed in  section 7.4 are adequate for
            this purpose, as shown in the following examples.


                                       27

-------
7.5.2  Elbow meters.
       7.5.2.1   The accelerations  associated with  flow  around  a
                curve of radius  r  cause  a  pressure difference  in the
                radial  direction which can be  used to deduce a
                flowrate (as well  as  monitor)  in a full  pipe.  This
                method, which has  been investigated mainly  for
                90-degree elbows,  requires an  arrangement as shown
                in figure 5.
                                                           A-A
                          Figure 5.  Elbow meter.
       7.5.2.2  The flowrate can  be  estimated  from the  following
                analytically determined  expression (14).

                         Q= (r/2D)1/2UD2/4)(29Ah)1/2       [4]

                Here,  r is  the  radius  of curvature of the elbow
                centerline,  D is  the elbow  (and  pipe) diameter and
                Ah is  the measured head  difference in terms  of height
                of flowing  liquid.                     ''
                                  28

-------
       7.5.2.3  The elbow performance is more sensitive to the radius
                of curvature of the inside bend than to that of the
                outside bend.  Therefore it is desirable, when
                practicable, to determine r by measuring the inner
                bend curvature and adding half the diameter rather
                than to use a nominal value of r.

       7.5.2.4  The elbow should be preceded by about 10 diameters of
                straight upstream pipe.   There is  insufficient
                information with which to evaluate such effects as
                pipe roughness and Reynolds number, except to note
                that with decreasing Reynolds number the flow is less
                than that predicted by equation [4].   Therefore added
                caution must be exercised in its application to sludge
                f1ows.

       7.5.2.5  Examination of published experimental  results suggests
                that equation [4] cannot be depended upon for accuracies
                better  than roughly +_ 10 percent.   However, it should
                be noted that an elbow meter that  is carefully fabricated
                and installed and properly calibrated can be as effective
                a flowmeter as other types of pressure-differential
                devices.  The foregoing  paragraphs of section 7.5.2
                pertain to uncalibrated  elbows only.

7.5.3  Valves.  Butterfly valves are sometimes furnished with flowrate
       vs.  angle-of-opening data which can be used for approximate
       checks on flowmeters.  Such measurements would of course be
       affected to an unknown extent by  upstream pipe conditions
       and  viscosity.
                                 29

-------
                         8.  OPERATION AND MAINTENANCE

8.1  Operation/Maintenance Frequencies

     8.1.1  Flowmeters without monitoring stations Csection 7.4).

            8.1.1.1  The frequency of required routine operation checks and
                     maintenance can best be determined from experience.
                     However, until experience provides guidelines,
                     relatively frequent checks must be made, particularly
                     where there is no monitoring station.
                     - Stop flow through the meter briefly at least once a
                       day to check for zero drift.  This should be done
                       initially even for meters that have internal zero
                       correction in order to check for possible offsets;
                       the frequency of subsequent checks can be reduced
                       as appropriate.
                     - Check the secondary span weekly.  See section 4.3.2.
                     - Visually inspect the liner and electrodes monthly
                       and clean as necessary.

            8.1.1.2  The frequency of the above checks can be adjusted as a
                     data base is developed.

     8.1.2  Flowmeters with monitoring stations.

            8.1.2.1  If there is a monitoring station as described in
                     section 7.4, the initial operation checks can consist
                     of a daily comparison of the electromagnetic flowmeter
                     reading with the monitor, provided that the satisfactory
                     condition of the internal surfaces and the pressure
                     taps of the monitor can be assured.  If this condition
                     is satisfied, the operations in  section 8.1.1.1  need
                     not be done at specified intervals until  the previously
                     determined relation (section 7.4.1.4) between the flow-
                     meter and monitor readings changes, at which time zero
                     and secondary span checks should be made, followed by
                     inspection and cleaning if necessary.

            8.1.2.2  If the internal  condition of the monitoring station cannot
                     be verified satisfactorily, treat the flowmeter according
                     to section 8.1.1  and use the monitor only to detect gross
                     malfunctions.
                                     30

-------
8.2  Miscellaneous Operation/Maintenance Recommendations

     8.2.1  Follow all  maintenance procedures recommended by the manufacturer
            in the user manual, in addition to those in section 8.1.

     8.2.2  If the flowmeter has been off line, always turn on the electrical
            supply several  hours before the readings are started.

     8.2.3  Measure the actual  internal diameter of the flow tube before
            it is installed.

     8.2.4  If the flowmeter is in a horizontal line (not recommended)
            vent the gas bleeds (section 5.3.2.2) at least once per shift
            until a more realistic frequency is learned from experience.
                                      31

-------
                                9.   REFERENCES

(1)   Theenhaus, R., "Modern Developments and New Applications of Magnetic
     Flowmeters," International Conference on Modern Developments in Flow
     Measurement, Atomic Energy Research Establishment, Harwell, G.  B.  1971,
     Paper No. 29.
(2)   Shercliff, J.  S., "The Theory of Electromagnetic Flow Measurement,"
     Cambridge University Press, 1962.
(3)   British Standards Institution, "Specification for Electromagnetic
     Flowmeters," BS5792:1980.
(4)   de Jong* J., "Comparison  of Some 500-mm Diameter Electromagnetic
     Flowmeters," in Flow Measurement of Fluids, H. H. Dijstelbergen
     and E.  A. Spencer, eds., North-Holland Publishing Co., 1978.
(5)   Scott, R. W. W.,  "A Practical Assessment of the Performance of Electro-
     magnetic Flowmeters," in Fluid Flow Measurement in the Mid 1970's,
     Vol. 1, edited  by  E.  A. Spencer and W. J. Ramsay, Nat. Eng. Lab.,
     G.  B.,  1977.
(6)   Haacke, A. C., "Calibration of Electromagnetic Flowmeters Using a
     Flow Simulator,"  in Flow Measurement of Fluids, H. H.  Dijstelbergen
     and E.  A. Spencer, eds.* North-Holland Publishing Co., 1978.
(7)   International  Standards Organization, "Measurement of Fluid Flow  by
     Means of Orifice Plates, Nozzles and Venturi  tubes Inserted in
     Circular Cross-Section Conduits Running Full," ISO/DIS 5167, 1976,
     draft revision  of R781.
(8)   American Society  of Mechanical Engineers,  "Fluid Meters  -- Their  Theory
     and Application,"  6th ed., 1971, 345 E. 47  St., New York, N.Y.  10017.
(9)   American Society  for Testing and MaterialSi  "Standard Method for  Open
     Channel Flow Measurement of Industrial Water and Industrial Waste  Water
     by the  PaYshall Flume," ASTM D1941-67.
(10)  British Standards Institution, Standard No.  2680-4A,  "Methods  of
      Measurement of Liquid Flow in Open Channels:   Part 4A, Thin Plate
      Weirs  and Venturi Flumes," 1965.
(11)  International  Standards Organization, "Measurement of Water Flow  in
      Closed Conduits—Tracer Methods,  Part I:  General," ISO  No. 2975/1,
      1974;  "Part II:   Constant Rate Injection Method Using Non-radioactive
      Tracers," ISO  No. 2975/2^ 1974.
(12)  Hydraulic Institute, "Standards for Centrifugal, Rotary  and Reciprocating
      Pumps," 12th edition.
(13)  International  Standards Organization, "Liquid Flow Measurement in
      Open Channels  —  Velocity Area Methods," ISO 748, 1976.
(14)  Replogle, J. A.,  et  al.,  "Evaluation of Pipe Elbows as Flow Meters,"
      Proc.  Am. Soc.  Civil Eng., Vol. 92, IR3, Sept. 1966,  17-34.
                                    32

-------