United States           Office of Water           EPA-No. TBD
           Environmental Protection         Washington, DC 20460      July 1998
           Agency


&EPA     Guide to Laboratory Contracting
           July 1998

-------
                                Acknowledgments
This document was prepared under the direction of William A. Telliard of the Engineering and
Analysis Division within EPA's Office of Water. It was originally developed to provide laboratory
contracting guidance to the Office of Wastewater Management's pretreatment program, but has
been modified to provide broader application to any organization requiring analysis of
environmental samples. This document was prepared under EPA Contract No. 68-C3-0337 by
DynCorp Information & Engineering Technology.
Questions concerning the guidance in this document should be addressed to:

       William A. Telliard
       Engineering and Analysis Division (4303)
       U.S. Environmental Protection Agency
       401 M Street, SW
       Washington, DC 20460
       Phone:(202)260-7120
       Fax:   (202)260-7185
Requests for additional copies of this document should be directed to:

      Water Resource Center
      Mail Code RC-4100
      401  M Street, SW
      Washington, DC 20460
      (202) 260-7786 or (202) 260-2814

-------
TABLE OF CONTENTS


1      Introduction	1

2      When Should You Outsource Analytical Services?	3

3      Developing an Analytical Contract	5
       3.1    Defining the Project Parameters 	5
             3.1.1  Client Information	5
             3.1.2  Number, Frequency, and Matrix of Samples  	5
             3.1.3  Project Background  	7
             3.1.4  Project Schedule and Data Turnaround Times	7
             3.1.5  Methodology  	7
       3.2    Selecting Appropriate Methods	8
             3.2.1  Methods Approved for Nationwide Use in Wastewater and Drinking WateiS
             3.2.2  Wastewater Methods Approved for Specific Industrial Categories	9
             3.2.3  Other Methods  	9
       3.3    Determining Appropriate Quality Control Requirements	10
             3.3.1  Established Laboratory Quality System	11
             3.3.2  Purity and Traceability of Reference Standards  	11
             3.3.3  Calibration Range	11
             3.3.4  Linearity of Calibration	12
             3.3.5  Calibration Verification	13
             3.3.6  Method Detection Limit, Minimum Level, or Quantitation Limit	13
             3.3.7  Initial Precision and Recovery	14
             3.3.8  Ongoing Precision and Recovery		... 14
             3.3.9  Analysis of Blanks	14
             3.3.10 Matrix Spikes and Labeled Compound Spikes	15
             3.3.11 Statements of Data Quality for Recovery of Spiked Analytes or
                    Labeled Compounds in Samples  	15
       3.4    Writing the Contract	16
             3.4.1  Deliverables  	17
             3.4.2  Data Turnaround Times  	17
             3.4.3  Liquidated Damages and Penalties  	18
             3.4.4  Reanalysis Costs	18
             3.4.5  Dilutions	19
       3.5    Developing a Bid  Sheet  	19
       3.6    Estimating Costs	20

4      Soliciting and Awarding the Contract	21
       4.1    Identifying Capable Laboratories and Transmitting the Requirements  	22
       4.2    Evaluating Bids  	22
       4.3    Conducting Responsibility Determinations  	23
             4.3.1  Method Performance Data  	23
             4.3.2  Performance Evaluation Sample Results  	24
             4.3.3  Certifications	25
             4.3.4  References 	25
                                           in

-------
              4.3.5  Prequalification Analyses  	26
       4.4    Awarding the Contract	28

5      Transporting Samples and Communicating with the Laboratory  	29
       5.1    Transporting and Tracking Samples  	29
       5.2    Communicating with the Laboratory	31

6      Reviewing Analytical Data	33
       6.1    Purity and Traceability of Reference Standards	33
       6.2    Calibration Range	33
       6.3    Linearity of Calibration	34
       6.4    Calibration Verification	34
       6.5    Method Detection Limit, Minimum Level, or Quantitation Limit  	34
       6.6    Initial Precision and Recovery	35
       6.7    Ongoing Precision and Recovery  	35
       6.8    Analysis of Blanks	 . .	 36
       6.9    Matrix Spikes and Labeled Compound Spikes  	37
       6.10   Statements of Data Quality for Recovery of Spiked Analytes or Labeled
              Compounds in Samples	38
       6.11   Field Duplicates 	38

7      References	41
Appendix A:  Analytical Services Request Form Example

Appendix B:  Bid Sheet Example

Appendix C:  General Laboratory Audit Checklist Example

Appendix D:  Data Review Checklist Example
                                           IV

-------
                                                                                Introduction
 1     INTRODUCTION
This document is intended to provide wastewater and drinking water treatment plants, states,
industrial facilities, and other parties requiring analysis of environmental samples with guidance on
laboratory contracting. It is based on laboratory contracting guidance developed specifically for
the Office of Wastewater Management for use with pretreatment authorities and industrial users.
This document provides guidance on the entire laboratory contracting process, from determining
when to use a contract laboratory rather than perform in-house analyses to development of the
analytical requirements and solicitation of the contract, to evaluation of laboratories and data
review.

Regardless of the reason for the analyses, or the nature of the organization requiring laboratory
services, the basic steps needed to ensure that reliable, useable data is generated in a timely
manner are the same. These steps are outlined in the box below.

     Critical Steps to a Successful  Laboratory Contract
       Step 1 - Clearly define your analytical needs
                 / Number and type of samples
                 / Applicable method(s)
                 / Quality control (QC) requirements
       Step 2 - Develop a clearly defined contract
                 / Analytical and QC requirements
                 / Data deliverables
                 / Data turnaround
                 / Contract enforcement clauses
       Step 3 - Solicit and award the contract openly and fairly
                 / Use standardized bid sheet
                 / Ensure laboratory is qualified before awarding the contract

       Step 4 - Maintain communications with the laboratory after sample receipt


       Step 5 - Thoroughly review the resulting data
Each of these general steps, as well as all of the other steps in between, are discussed in detail in
this guide. By following the guidance in this document, you can obtain the data you need, when
you need it, at reasonable costs and with minimal analytical problems.

-------
                                                  When Should You Outsource Analytical Services?
       WHEN SHOULD You OUTSOURCE ANALYTICAL SERVICES?
A fundamental decision when planning for laboratory analyses is whether the analyses should be
performed by an in-house laboratory or whether the project should be contracted out to a
commercial laboratory. Two factors impact whether the in-house laboratory can accept an
analytical project: capacity and capability. An in-house laboratory must have: (1) the capacity to
analyze the quantity of samples requested within the required time period, (2) the instrumentation
expertise to perform the required analyses, and (3) staff qualified to perform the analyses.

If your laboratory does not have the capacity or the capability to handle the analytical
requirements, you must decide whether it is appropriate to outsource the analyses or to increase
the capacity or capability of the in-house  laboratory. This decision is a function of time and cost.

The time required to increase the laboratory's capacity or capability will include the time required
to identify and obtain the necessary equipment and instrumentation, the time necessary to bring
instrumentation on-line, the time necessary to train staff in new procedures, and the time required
to perform any necessary start-up tests and demonstrations of analyst capability. Each of these
must be completed prior to project deadlines.

Determining the cost-effectiveness of increasing capacity or developing additional capability is a
complex issue, however, the equation below provides a framework for this decision:
                                                   T) +R + U)O
                                         TV
Where:

  I  =  cost of required instrumentation

  5  =  years before instrument depreciation reaches 100%

  FJ =  square footage required for instrument and work area

  Ft =  square-foot cost of total floor space per year

  A =  number of analysts required to perform the analysis

  S  =  annual salary of analysts

  P  =  percentage of each analyst's time dedicated to the new capability (for example, 0.80)
  C =  number of hours of calibration and maintenance required per year

  T =  hourly cost of technician for maintenance, if not the primary analyst

-------
Guidance on Procuring Analytical Services for Environmental Samples
  R  = annual cost of required reagents and other supplies, such as pressurized gas or liquid
       nitrogen

  U  = increase in annual utility cost to power the instrument and provide air conditioning

  O  = overhead factor for management (generally 1.2 - 2.0)

  N  = number of analyses per year anticipated over five years

Estimates for the annual number of hours of calibration and maintenance time (C) typically can be
provided by instrument vendors or in-house staff with previous experience operating these
instruments. In some cases, facilities may have instrumentation that is no longer in use, but may be
serviceable. In such cases, the cost of required instrumentation (I)  should be omitted from the
equation. Similarly, the square footage required for the instrument and work area (F J and the cost
per square foot of total floor space per year (Ft) should be modified to reflect only the added work
area  space needed to operate the already-housed equipment.

If x ^ the average per-sample analysis cost at a. commercial laboratory for the same analysis, you
should consider expanding the your in-house laboratory capability accordingly, and begin
analyzing such samples in-house. However, if x > the average per-sample analysis cost at a
commercial laboratory, you should consider contracting the analyses with a commercial
laboratory.  This guidance manual addresses the subsequent steps that should be taken to ensure
delivery of quality data after the decision has been made to out-source the sample.

-------
                                                             Developing an Analytical Contract
       DEVELOPING AN ANALYTICAL CONTRACT
Although most organizations have established procedures and policies governing the purchase of
services and supplies, these procedures seldom lend themselves to the purchase of analytical
services, primarily due to the difficulty in defining the required services. This chapter provides a
basic framework for defining the technical and contractual requirements associated with
purchasing analytical services.

3.1    Defining the Project  Parameters

Many laboratories have recognized the importance of customer service and employ staff who are
trained to assist clients in defining specific requirements that meet their needs. Other laboratories,
large and small, rely solely on the client to define these requirements. Still another group of
laboratories, albeit a small group, perform analyses with little regard to the client's actual needs.
One of the problems that arises when the client's requirements are poorly  defined is the use of
inappropriate methods. As detailed in Section 3.2, approved methods must be used for
compliance monitoring purposes. It  is not the laboratory's place to decide that an alternate
method, even if it is an alternate EPA method, is "close enough."  You are responsible for
defining and ensuring that contract laboratories adhere to requirements that are consistent with
your monitoring requirements. The first step in developing an analytical services contract is
identifying the who, what, why, when, and how of the project. The remainder of this subsection
provides guidance for defining these project parameters.

3.1.1  Client Information

Who is the name of the client, whether this is a single wastewater or drinking water plant or
industrial facility, a utility or company that oversees several plants, or a third-party organization
interested in generating original data. Client information should include specific contact points for
use by the laboratory, including the following:
       The name of the person responsible for communicating with the laboratory regarding
       shipping delays and broken samples (a sample control contact)

       A technical contact for resolution of analytical questions or problems (a technical contact)

•       An administrative contact for invoicing and payment (a billing contact)

Include with the contact name their address, telephone and fax numbers, and email,  as these
contacts often are in different locations.

3.1.2  Number, Frequency, and Matrix of Samples

What describes the samples to be analyzed, including:
•      Number of "billable" samples and number of "unbillable samples." Billable samples
      (sample analyses for  which the laboratory will be paid their per-sample cost) typically

-------
Guidance on Procuring Analytical Services for Environmental Samples
       include field samples, while unbillable samples (sample analyses for which the laboratory
       will not be paid) typically include internal laboratory quality control (QC) samples, such as
       blanks and ongoing precision and recovery (OPR) samples. Because the billable nature of
       some samples, such as matrix spike/matrix spike duplicate (MS/MSD) samples and field
       blanks is addressed differently by different organizations, you must clearly define which
       types of samples will be billable and which types will be unbillable, and provide the number
       of each type anticipated.

•      Frequency with which samples will be sent to the laboratory (for example, five samples per
       week for eight weeks). This is particularly important for long-term projects, as it allows
       laboratories to accurately evaluate whether they have the capacity to analyze the samples
       and may result in volume discounts.
•      Analyses required (for example, volatile organics, pesticides, or metals). Methods at 40
       CFR Part  136 must be used for any National Pollutant Discharge Elimination System
       (NPDES) analyses unless otherwise specified in your NPDES permit. Drinking water
       utilities must use methods listed at 40 CFR Part 141.

       Sample matrices (for example, raw source water or finished drinking water, wastewater,
       sludge, or solids)

The number of samples, frequency of collection, and the types of analyses required are relatively
straightforward issues, if properly defined, as noted above. However, special care should be taken
to accurately and thoroughly describe the sample matrix because laboratories often attribute the
inability to measure the concentration of a pollutant in a specific wastewater to "matrix
problems."  Most matrix interferences can be overcome with sufficient time, equipment, and
procedures. Therefore, it is important to provide laboratory staff with as much advance
information as possible concerning sample matrices so that the staff can adequately plan for
analysis of any complex samples and avoid delays, unanticipated cost increases, or the generation
of unusable data after samples are collected and shipped.

Examples of potential matrix problems include samples that contain high levels of organic
compounds and samples that appear to be biphasic. Generally, the laboratory analyst should be
responsible for evaluating specific matrix problems, developing solutions to matrix interference
problems, and presenting recommended solutions to the client for approval. If, however, the client
already is aware of matrix interferences and solutions that have  been demonstrated to remove
these interferences, the client should provide this information to the laboratory during project
solicitation or, at the latest, when the samples are scheduled.

For compliance monitoring purposes, permittees must ensure that any modifications they make to
improve method performance are legally acceptable. Most NPDES permits require the use of
methods approved at 40 CFR Part  136, while drinking water monitoring requires the use of
methods approved at 40 CFR Part 141. Many methods provide the analyst with the flexibility to
modify the method as the analyst demonstrates that the method modifications  produce results that
are as good or better than those produced by the original method. If the method required for
monitoring does not include such flexibility, it is incumbent on you to seek written permission for
any and all method modifications from your permitting authority, state, or EPA Region. In doing
so, you should work with the analytical laboratory and the permitting authority, state, or EPA

-------
                                                              Developing an Analytical Contract
Region to identify the requirements necessary for demonstrating that the method modifications are
appropriate.

The costs associated with removing matrix interferences vary from laboratory to laboratory and
problem to problem, as do the costs of the basic analysis. The costs involved in modifying a given
method to overcome a complex matrix problem and in validating the use of additional cleanup
techniques could range between several hundred and several thousand dollars, depending on the
complexity of the wastewater, the experience of the laboratory in resolving matrix interferences,
and the flexibility of the method. These costs associated with known matrix problems should be
addressed with the laboratories' bids.

3.1.3   Project Background

Why often is overlooked because personnel assume that all parties  involved understand the
purpose of the analysis. Providing information on why the analyses are required, such as stating
that wastewater samples are to be analyzed for NPDES permit compliance monitoring or that
samples are being analyzed by various methods to characterize a waste stream, will help the
laboratory provide the data you want.

3.1.4   Project Schedule and Data Turnaround Times

When specifies the following dates:

•      The date by which bids should be received  from laboratories interested in the project (this
       is discussed in greater detail in Section 3.5)
•      The approximate date that the samples will  be shipped to the laboratory, including the
       means of shipment (such as overnight delivery, hand-delivery, or laboratory pick-up)

•      The date the analytical results are required by the client (data turnaround time) (this is
       discussed in greater detail in Section 3.4.2 and 3.4.3)

The data turnaround time specifies the number of calendar days after the laboratory receives the
sample that the results are to be received by the client. Common data turnaround times are 30 to
45 days after the laboratory receives the last sample. The data turnaround time can be specified on
a method-specific basis, and often is a function of a reporting deadline under a permit. Because
commercial laboratories typically operate in production mode, where samples are processed
through a  standard sequence of login, preparation, digestion/extraction, analysis, data production,
and quality assurance (QA), you can reduce analytical costs by providing the laboratory with as
much time as possible to provide the data. In addition, some laboratories may be willing to
negotiate a discounted fee if you promise to accelerate payment within a certain number of days.
Caution should be exercised with this option, however, as it may prevent you from thoroughly
reviewing the quality of the data prior to payment. Turnaround times extending beyond 45 to 60
days rarely provides any significant cost savings.

3.1.5  Methodology

How is perhaps the most important question to be addressed in the  analytical services contract,
and specifies the required methodology, the quality assurance/quality control (QA/QC), and the
reporting format. The analytical requirements must  be very specific, and include the following:

-------
 Guidance on Procuring Analytical Services for Environmental Samples
       Method source and number. Methods approved for wastewater analyses are listed in
       Tables IA through ID at 40 CFR Part 136. Methods approved for drinking water analyses
       are listed at 40 CFR Part 141. Common method sources for wastewater and drinking
       water analyses include EPA methods, Standard Methods, American Society for Testing
       and Materials (ASTM) methods, and U.S. Geological Survey (USGS) methods. Other
       relevant method sources include EPA's solid waste methods (SW-846) and Association of
       Official Analytical Chemists (AOAC) methods. Selection of appropriate methods is
       discussed in detail in Section 3.2.

•      Holding time. The holding time of a sample is the maximum amount of time that can
       elapse between sample collection and analysis, between sample collection and extraction,
       or between sample extraction and analysis, for a sample result to be considered valid. The
       sample holding time for each method is different, and some methods may have more than
       one holding time. The holding times always should be specified. Table II at 40 CFR Part
       136 specifies the "Required Containers, Preservation Techniques, and Holding Times" for
       routine parameters. Samples should be analyzed as soon as possible after collection.
       Samples for many analyses are not stable  for long after collection, and daily shipment of
       samples to the laboratory should be considered. Delays in sampling and sample shipment
       may necessitate specifying a "contract" holding time in the contract, based on the
       analytical holding time minus any time required for sample shipment.

       Quality assurance/quality control. To ensure that data are valid and not a result of
       contamination or improperly calibrated instruments, laboratories must take rigorous
       QA/QC steps when performing the analyses. Specific guidance for defining QA/QC
       requirements is provided in Section 3.3. To ensure that the proper QA/QC steps are being
       routinely followed, all data must be reviewed and validated, and the laboratory should be
       audited at a frequency commensurate with the length, scope, and importance of the
       project.

       Delivcrablcs and reporting format. The laboratory also needs to know how the data are
       to be reported, what information to provide in addition to the results, and how many
       copies of the data package are required. Deliverables are discussed in greater detail in
       Section 3.4.1.

3.2   Selecting Appropriate  Methods

The methods approved by EPA for nationwide use under the Clean Water Act (CWA) and Safe
Drinking Water Act will satisfy most of your analytical needs. However, some industrial
wastewaters may cause matrix-specific analytical  problems, while other matrices may contain
analytes of interest for which there are no EPA-approved methods. Sections 3.2.1 - 3.2.3 present
information on methods generally applicable to analysis of wastewater and drinking water
matrices. Exceptions to the information presented below may arise in specific permit situations;
therefore, all permittees must be familiar with the terms and requirements of their individual
permits.

3.2.1   Methods Approved for Nationwide  Use in Wastewater and Drinking Water

EPA publishes test procedures for measuring regulated contaminants in wastewater and drinking
water at 40 CFR Part 136 and 40 CFR Part 141, respectively. The methods approved for use at

-------
                                                             Developing an Analytical Contract
40 CFR Parts 136 and 141 include EPA methods, Standard Methods, ASTM methods, USGS
methods, and Department of Energy methods. Other resources for analytical methods can be
found on the EPA Office of Water's Environmental Monitoring Methods Index database
(Reference 1) and Methods and Guidance for Analysis of Water CD-ROM (Reference 2). These
resources allow the user to search for methods by analyte, and provide information on detection
limits, instrumentation, and applicable matrices. Additional information on applicable regulations
and method guidance also can be found on these resources.

Wastewater plants should be aware that the results of any final effluent analyses that are
conducted with test procedures approved at 40 CFR Part 136 must be reported with the data
submitted in the permit-required monitoring report, even if those analyses were not required in the
permit. Results of analyses  conducted with non-approved methods or conducted on unregulated
waste streams are not required to be reported (40 CFR Part 122.41).

The 40 CFR Part 136 methods are applicable to a wide range of industrial effluents and typically
were used to generate the data necessary for developing the effluent guidelines promulgated by
EPA. Despite this wide applicability, EPA recognizes that these analytical methods may fail to
yield useful results when used on certain sample matrices. EPA is prepared to consider claims that
the effluent is compliant in those instances in which the effects of the sample matrix make
measurements difficult or impossible. However, all such claims must be supported by specific
analytical data that demonstrates reasonable, but unsuccessful, attempts have been made to
overcome matrix interferences.

3.2.2  Wastewater Methods Approved for Specific Industrial Categories

In addition to methods approved at 40 CFR Part  136 for all wastewaters, EPA promulgates
methods to measure pollutants for specific industrial categories at 40 CFR Parts 405 - 471, along
with that industry's categorical effluent limitations and guidelines. Because EPA's authority to
promulgate wastewater methods is specified at Section 304(h) of the Clean Water Act, methods
promulgated at 40 CFR Part 136 or at 40 CFR Parts 405 - 471 often are referred to as the
"304(h) methods."

3.2.3  Other Methods

If no approved methods are applicable to the analytes of interest or the matrix, selection of non-
approved methods may be necessary (note, however, that unapproved methods cannot be used
when a comparable approved method is available unless a permit or state or EPA Region
explicitly allows the use of such alternate methods). In  such instances, appropriate QA/QC
procedures must be performed and low level detection limits must be achievable as necessary to
demonstrate compliance with applicable permit limits or maximum contaminant levels (MCLs). As
mentioned above, the permitting authority must approve the use of these methods  in advance.

3.2.4  Method Modifications

Many of the methods approved at 40 CFR Parts 136 and  141 provide flexibility to modify the
method to improve method performance, reduce  cost, or adapt the method  to address more
difficult matrices without prior approval. Example improvements include the use of additional
cleanup techniques, alternative gas chromatography or liquid chromatography columns, and more

-------
Guidance on Procuring Analytical Services for Environmental Samples
specific detectors. However, the modifications to these methods cannot result in any degradation
of method performance, and the laboratory that analyzes the samples with a modified method
must first demonstrate that the modifications result in performance equivalent to that of the
reference method. At present, only some methods approved at 40 CFR Parts 136 and 141 provide
this flexibility. Unless this flexibility is provided explicitly in the method, modifications to a
methods must be approved through the Alternate Test Procedure program.

In October 1995, EPA proposed the use of several new and modified methods for monitoring
inorganic pollutants at 60 FR 53988. Most of the methods included in that proposed rule provide
laboratories with the flexibility described above. In March 1997, EPA proposed to expand and
extend this flexibility to nearly all methods approved for use at 40 CFR Part 136 and 40 CFR Part
141. Details of this proposal, known as the streamlining initiative, are given at 62 FR 14976.

If methods approved at 40 CFR Parts 136 or 141 fail to yield acceptable results in a specific
matrix, you should consider modifying an approved method to yield improved and acceptable
performance.  Similarly, if methods approved at 40 CFR Part 136 of 141 are not applicable to a
specific wastewater or drinking water pollutant that you need to monitor, approved methods
applicable to similar pollutants may exist. In such cases, these entities may wish to consider
modifying an approved method to target the pollutant of interest. In all cases, the modifications
must either be allowed in the approved version of the method through the flexibility described
above, or must be approved by the permitting authority or state or EPA Region. Permittees
seeking approval of method modifications are encouraged to use the streamlining proposal at 62
FR 14976 as a basis for initiating discussions with their permitting authority.

3.3   Determining Appropriate Quality Control Requirements

The ability to  define the validity of your sample results is wholly based on the degree of QC data
associated with your results. Therefore, you are strongly encouraged to use the guidelines
provided in this section, or similarly developed standard protocols, to establish strict data quality
requirements for the analyses performed by contract laboratories (use of these requirements for
in-house analyses also is advised). These QA/QC requirements subsequently enable the end-users
of the data to standardize data inspection and acceptance procedures. This minimizes differences
that might otherwise result between data reviewers and laboratories.

A standardized QA/QC approach should take the form of performance  specifications for each
method and should contain the following elements:

•       Established laboratory quality system

•       Purity and traceability of reference standards

       Calibration Range
       Linearity of calibration

       Calibration verification

•       Method detection limit (MDL), minimum level (ML), or quantitation limit

•       Initial  precision and recovery (IPR)

                                           10

-------
                                                              Developing an Analytical Contract
•      Ongoing precision and recovery (OPR)

       Analysis of blanks

•      Recovery of matrix spikes and labeled compound spikes

       Statements of data quality for recovery of spiked analytes or labeled compounds in
       samples
       Analysis of field duplicates

These elements are an integral part of many recent EPA methods. However, earlier methods may
not specify some or all of these elements. As such, you should ensure that any QA/QC elements
not specified in the method(s) required under the contract are specified in the contract itself. A
summary of each of the elements is provided below. Guidance on assessing data using these
QA/QC results is provided in Section 6. The analytical laboratory should be audited to ensure that
QA/QC procedures are being implemented on a daily basis. The frequency of these audits should
be commensurate with the length, scope, and importance of the project. Audit information on a
laboratory may be obtained by contacting the QA Officers or inspection staff in state and EPA
regional offices with jurisdiction.

3.3.1  Established Laboratory Quality System

Any laboratory that performs analyses to support permitting compliance and monitoring should be
required to have an established quality system that is compliant with ISO/IEC Guide 25: General
Requirements for the Competence of Calibration and Testing Laboratories (Reference 3). This
document sets forth general QA/QC guidelines for laboratories to follow, including personnel,
analysis environment, and equipment requirements, requirements for internal reviews and audits,
and the other requirements specified in Sections 3.3.2 - 3.3.11. It is essential for laboratories to
employ comprehensive quality systems throughout the duration of the contract to ensure data
validity. The laboratory should implement the quality system, and should otherwise use safe
handling procedures and employ accepted Good Laboratory Practices in all aspects of laboratory
performance.

3.3.2  Purity and Traceability of Reference Standards

The accuracy of any non-absolute empirical measurement depends on the reference for that
measurement. In determining pollutants in water or other sample matrices, laboratories must
calibrate analytical instruments and analytical processes with a known reference material. Most of
the methods approved at 40 CFR Parts 136 and 141 require that the standards used for calibration
and other purposes be of known purity and be traceable to a reliable reference source.  The
ultimate source for reference materials is typically EPA or the National Institute for Standards and
Technology (NIST).

3.3.3  Calibration Range

Instrument calibration  is required to establish the relationship of analyte concentration to
instrument response, and is subsequently used for the quantitative analysis of field samples. This
relationship is determined by analyzing a series of reference standards at different concentration
                                           11

-------
Guidance on Procuring Analytical Services for Environmental Samples
levels (calibration points) which encompass the expected concentration range of field samples and
the expected linear range of the analytical instrument.

Most EPA methods for organic pollutants specify a minimum of three calibration points. Newer
methods for inorganic pollutants also specify a minimum of three calibration points. The lowest of
these points is required to be at or near the MDL. The highest is required to be near the upper
linear range of the analytical system, and the third point is approximately midway between the
two. The lowest calibration point  should never be greater than five times the MDL and should
ideally be within three times the MDL. The results for the lowest calibration standard are the
principal means by which to assure that measurements at levels near the MDL are reliable. The
EPA Office of Water uses the lowest calibration standard as one means of defining the ML of
quantitation.

The flexibility in selecting the levels of the calibration points in many EPA methods has led to a
wide variety of calibration ranges  as each laboratory may determine its own calibration range.
Some laboratories establish a relatively narrow calibration range, such as a five-fold increase in
concentration, because it makes it simpler to meet the linearity specifications of the method. Other
laboratories choose wider calibration ranges in order to minimize the number of samples that have
to be diluted and reanalyzed because the concentration of one or more analytes exceeds the
calibration range. Understanding these differences is particularly important if a narrow
concentration range results in increased costs of sample dilution or if the laboratory's
concentration range prevents the laboratories from achieving the required detection or
quantitation limits.

3.3.4  Linearity of Calibration

The relationship between the response of an analytical instrument to the concentration or amount
of an analyte introduced into the instrument is referred to as the "calibration curve." An analytical
instrument can be said to be calibrated when this relationship has been established.  The ratio of
the response of the instrument to the concentration of the analyte introduced into the instrument is
called the response factor (RF), relative response factor (RR), or calibration factor (CF):

•      Relative response (RR) for isotope dilution calibration

•      Response factor (RF) for internal standard calibration

•      Calibration factor (CF) for external standard calibration

A plot of instrument response and concentrations is generated, and the linearity of response is
measured by the shape of the calibration curve. While the shape of calibration curves can be
modeled by quadratic equations or higher order mathematical functions, most analytical methods
recommend establishing a linear calibration. The advantage of the linear calibration is that the RF
or RR represents the slope of calibration curve, simplifying calculations and data interpretation.
The 1600 Series Analysis Methods contain specific criteria for determining the linearity of
calibration curves determined by either an internal or external standard technique. When the
applicable criterion is met, the calibration curve is sufficiently linear to permit the laboratory to
use an average RF or RR, and it is assumed that the calibration curve is a straight line that passes
through the zero/zero calibration point. Linearity is determined by calculating the relative standard
deviation (RSD) of the RF or RR  for each analyte and comparing this RSD to the specified limit.

                                             12

-------
                                                                Developing an Analytical Contract
The number of calibration points is dependent on the error of the measuring technique.
Measurement technique error is determined by (1) calibrating the instrument at the ML of
quantitation and a minimum of two additional points, and (2) determining the RSD of the RR, RF,
or CF. For most analyses, such as the determination of semi-volatile organic compounds by
extraction, concentration, and gas chromatography, the measuring instrument is calibrated, and
sample preparation processes are excluded from the calibration process; for others, such as the
determination of purgeable organic compounds by purge-and-trap gas chromatography,
calibration encompasses the entire  analytical process. Table 3-1 below gives the number of
calibration points required depending on the calibration linearity.

Table 3-1.     Minimum Number of Points Required for Calibration
Percent RSD1
0-<2
2-<10
10-<25
>25
Minimum Number of Calibration Points
12
3
5
7
'Percent RSD shall be determined from the calibration linearity test for replicate measurements at a fixed concentration.
2Assumes linearity through the origin (0,0). For analytes for which there is no origin (such as pH), a two-point calibration shall be
performed. In almost no cases should only one calibration point be used. One calibration point most often leads to serious error.

The maximum RSD specification is applicable to calibration with three or more calibration points.
Alternatively, a minimum correlation coefficient for the linear relationship may be specified, below
which the calibration linearity is not acceptable. If the calibration curve is non-linear, a second
order (y = ax2 + bx + c) calibration curve may be used. Calibration functions higher than the
second order are not allowed.

3.3.5  Calibration Verification

Calibration verification involves the analysis of a single standard, typically in the middle of the
calibration range, at the beginning  (and in some cases, at the end) of each analytical shift. The
concentration of each analyte in the reference standard is determined using the initial calibration
curve, and the results are compared with method specifications. This test is used to periodically
verify that instrument performance has not changed significantly. Specifications for calibration
verification are developed to define the allowable deviation of the RR, RF, or CF of the
calibration verification standard from the mean RR, RF, or CF of the initial calibration; or in cases
where the initial calibration curve did not meet linearity specifications, deviation from a prior
calibration verification standard or a single point of the calibration curve.

3.3.6  Method Detection Limit, Minimum Level, or Quantitation Limit

The Minimum Level (ML) is defined as the lowest level at which the entire analytical system gives
a recognizable signal and, in most instances, an acceptable calibration point. Procedures for
determining an MDL are provided at 40 CFR Part 136, Appendix B. Most of the 40 CFR Part
136, Appendix A and 40 CFR Part 141 Subpart C methods  contain MDLs, although few of the
methods explicitly require laboratories to demonstrate their ability to achieve these MDLs.
Laboratories that wish to practice any method on a routine basis should be required to
demonstrate that they can measure pollutants at the MDL or the detection limit specified in the


                                             13

-------
Guidance on Procuring Analytical Services for Environmental Samples
method. Performance of an MDL study in accordance with the 40 CFR Part 136, Appendix B,
procedure is one means of demonstrating such proficiency.

3.3.7  Initial Precision and Recovery

The IPR test is used as an initial demonstration of a laboratory's capability to produce results at
least as precise and accurate as those of other laboratories. The IPR test is also used to
demonstrate that a method modification will produce results as precise and accurate as results
produced by the  approved (reference) method. The IPR test consists of four aliquots of reagent
water spiked with the analytes of interest and with either surrogate compounds, or for isotope
dilution analysis, with the labeled compounds. The spike concentration of the target analytes in
the spike solution may vary between one and five times the lowest concentration used to establish
the calibration curve (such as one to five times the ML). The spiked aliquots are carried through
the entire analytical process. The mean concentration (x) and the standard deviation (s) are
calculated for each analyte and compared to the specifications  in the method. The IPR test is
performed by the laboratory before it uses a method or a method modification for analysis of
actual field samples.

3.3.8  Ongoing Precision and Recovery

The OPR test, sometimes termed a "laboratory control sample," "quality control check sample,"
or "laboratory-fortified blank," is used to ensure that the laboratory remains in control during the
period that samples are analyzed, and  separates laboratory performance from method performance
on the  sample matrix. The test consists of a single aliquot of reagent water spiked with the
analyte(s) of interest, which is carried  through the entire analytical process with each batch of
samples. Typically, the concentration of the target analyte(s) in the OPR sample is between one
and five times the lowest concentration used in the calibration curve (such as one to five times the
ML). The results of the OPR are compared with method specifications.

3.3.9  Analysis of Blanks

Blanks are analyzed either periodically or with each sample batch, and are analyzed to
demonstrate that no contamination is present that would affect the analysis of standards and
samples for the analytes of interest. Different types of blanks are analyzed to more precisely
determine if and  when contamination was introduced. The following are different types of blanks
that may be required by the methods selected for analysis:

       Initial and continuing calibration blanks (ICB/CCB). These blanks are required for all
       calibrated instrumentation.  Deionized distilled water that contains the same reagents as the
       prepared  samples is analyzed after analysis of the calibration standard to demonstrate the
       absence of carryover from the  standard into the sample.

       Preparation blanks. Deionized distilled water is carried through preparation and analysis,
       using the same sample preparation, reagents, and analysis methods used for field samples.
       Preparation blanks are prepared and analyzed with each sample set to demonstrate that
       contamination is not introduced during any of the sample preparation or analysis steps.

•       Blanks. Blanks are required for titrimetric and gravimetric methods, and any other method
       which does not require instrument calibration or sample preparation. Deionized distilled
       water which is not prepared, but contains the same reagents as the prepared field samples,

                                            14

-------
                                                             Developing an Analytical Contract
       is analyzed to determine if the method analyte or other interferences are present in the
       laboratory environment, reagents, or apparatus.

       Trip blanks. Trip blanks are generated by the sampler for volatile compounds and low
       level metals, such as mercury. These blanks consist of vials of water that accompany each
       sample shipment to determine whether contamination has occurred from permeation of
       volatile organic compounds or low-level metals during sample transportation.

       Equipment blanks. These blanks are sampler generated to determine contamination from
       compositor sampling line or tubing.

The types of blanks required for analysis is dependent on the requirements of each method, and
the period or batch size for which these blanks are required is also defined in each method. QC
acceptance  criteria are given in most methods. Generally, the source of contamination in a blank
analysis must be identified and eliminated before the analysis of standards and samples may begin.
Samples analyzed with an associated contaminated blank must be reanalyzed and, for
contaminated preparation blanks, reprepared.

3.3.10 Matrix Spikes and Labeled Compound Spikes

The non-isotope dilution methods require that laboratories spike the analytes of interest into a
second aliquot of a field sample and analyze this spiked sample with the non-spiked field sample.
The purpose of spiking the sample (often termed a matrix spike) is to determine if the method is
applicable to the sample matrix in question.  Most EPA methods were developed for the analysis
of wastewater effluent or treated drinking water samples, and may not be appropriate for in-
process samples. While many wastewater methods were tested using effluents from a wide variety
of industries, samples from some sources may not yield acceptable results. It is therefore
important to evaluate method performance in the sample matrix of interest.

If the recovery of the matrix spike is within the limits specified in the method, then the method is
judged to be applicable to that sample matrix. If, however, the recovery of the spike is not within
the recovery range specified, either the method does not work on the sample, or the sample
preparation process is out of control.  If the method is not appropriate for the sample matrix, then
changes to the method are required. Matrix spike results are necessary in evaluating the modified
method. If the analytical process is out of control, the laboratory must take immediate corrective
action before any more samples are analyzed.

To separate indications of method performance from those of laboratory performance, the
laboratory should prepare and analyze a QC check standard consisting of a spike of the analytes in
reagent water. If the results for the QC standard are not within the range specified, then the
analytical system must be repaired and the sample and spiked sample analyses repeated. If the
recovery of this spike is within the range specified, then the analytical process is judged to be in
control.

3.3.11 Statements of Data Quality for  Recovery of Spiked Analytes or Labeled
       Compounds  in Samples

EPA methods specify that after the analyses  of five spiked samples, a statement of data quality is
constructed for each analyte. The statement  of data quality for each analyte is computed as the

                                           15

-------
Guidance on Procuring Analytical Services for Environmental Samples
mean percent recovery plus and minus two times the standard deviation of percent recovery for
each analyte. The statements of data quality should then be updated by the laboratory after each
five to ten subsequent spiked sample analyses.

For non-isotope dilution results, the statement of data quality can be used to estimate the true
value of a reported result and to construct confidence bounds around the result. For example, if
the result reported for analysis of phenol is 25 /ug/L, and the statement of data quality for phenol
is 70% ± 15% (i.e., the mean recovery is 70% and the standard deviation of the recovery is 15%),
the true value for phenol will be in the range of 28 - 43 //g/L, with 95% confidence. This range is
derived as follows:

       Lower limit = [(25-0.7) - (25 x 0.3)]  =  [35.7 - 7.5]  =  28 //g/L
       Upper  limit  = [(25 - 0.7) + (25 x 0.3)] =  [35.7 + 7.5] = 43  //g/L

Statements of data quality for isotope dilution methods are based on the recoveries of the labeled
compounds. Using an isotope dilution method, the sample result has already been corrected for
the recovery of the labeled analog of the compound. Therefore, for a reported result for phenol of
25 Aig/L where the standard deviation of the labeled phenol recovery is 15%, the true value for
phenol will be in the range of 21.25-28.75 //g/L, with 95% confidence, derived as follows:

       Lower limit = [25 - (25 x  0.15)] =21.25 //g/L
       Upper  limit  = [25 + (25 x 0.15)]  = 28.75 /^g/L


3.4   Writing the Contract

Before writing a contract for any analytical services, consult with appropriate legal staff.  A well-
written contract will include the who, what, why, when, how issues outlined in Section 3.1, above.
It also will address your right to review the data as needed, the timeliness of payment to the
laboratory, and your ultimate right to determine that the work does not meet the requirements
established in the contract. A general format for an analytical services contract is provided in
Appendix A. Please note that the information  requested in Appendices A and B may not be
adequate for competitive, written solicitations to multiple laboratories; depending on the project,
more information may need to be requested in order to ensure the laboratory will be able to meet
the requirements of the analytical contract.

The best way to ensure that you get the required data within the required time period is to specify
these requirements in detail in the contract. Combined with a careful analysis of the requirements
discussed in Sections 3.1 through 3.3, a well-written contract can minimize or eliminate many
common problems in procuring analytical services. It should enable the client to obtain technically
sound, legally defensible, and timely analytical data to meet a variety of compliance monitoring
needs. Once generated, the basic form of the contract should be viewed as a dynamic document
that is routinely updated to clarify ambiguities that arise during its implementation. (Note: Active
contracts typically require a formal contract modification that is approved by both sides before its
terms can be changed; expired or closed contracts can be modified before they are re-issued.)

General issues that should be specified in the  contract are detailed in Sections 3.4.1 through 3.4.5.
                                            16

-------
                                                               Developing an Analytical Contract
3.4.1  Deliverables

You must ensure that the laboratory provides data that can be easily reviewed and that includes
non-quantitative information related to the analyses, such as descriptions of any problems
encountered. Laboratories should be required to have the following data from samples analyzed
available for review:

•      Summary reports of all analytical results in hardcopy and electronic data format. The
       summary report should contain a summary of analytical results for all QC and field
       samples. For the IPR analysis, the spiking level, individual results of the four replicates,
       and the mean recovery and relative standard deviation of the four replicates should be
       reported. For the OPR, standard reference material (SRM)/quality control sample (QCS),
       and calibration verification analyses, the true (or expected) concentration of the QC
       sample, the measured concentration, and the percent recovery should be reported. For
       MS/MSD analyses, the background concentration of the field sample, the spiking level, the
       individual results of the MS and MSD analyses, the percent recovery for the MS and
       MSD, the average concentration found in the MS/MSD samples, and the RPD between
       the MS and MSD should be reported. The results for all other QC, including calibration
       and blanks, also should be reported.

       A list of the sample numbers analyzed and a run chronology.

•      Copies of all raw data, including quantitation reports, strip charts, spectra, bench sheets
       and laboratory notebooks showing tare and sample weights, sample volumes, and other
       data that will allow the final results reported to be traced back to the analytical steps
       performed. Each data element should be clearly identified in the laboratory's data package.
•      A written report that details any problems associated with the analysis of the samples.

•      A detailed written description  of any approved modifications to the  procedures specified
       in the contract-specified method.

With the possible exception of electronically formatted data,  EPA recommends that you require
all of the above deliverables as part of the data submission by your contract laboratory(ies).

3.4.2  Data Turnaround Times

The required data turnaround must be stated clearly in the contract. Unless you can guarantee to
the laboratory that the samples will arrive when the laboratory opens in the morning, the data
turnaround time  calculations should consider the day that the sample is received at the laboratory
"day zero," and the following day as "day one." In addition to stating the time that the laboratory
has to generate and deliver the data, it may be useful to assign some specific consequences to the
possibility of late delivery. One approach is to assess a penalty of some percentage of the
analytical price per day of lateness. In the past, EPA has  used values of 1% or 2% per day after
the due date that the data were delivered. Obviously, lateness penalties should not be assessed if
the delays were due to changes in the requirements made after the samples  were sent, or to the
fact that the  methods  requested were not applicable to the samples. Many of the remedies to
matrix problems cannot be expected to be carried out in the original turnaround time assigned to
the sample unless those remedies were explicitly detailed and required in the contract (see Section
3.1.2). However, after you have established that your samples can routinely be analyzed by the
                                            17

-------
 Guidance on Procuring Analytical Services for Environmental Samples
requested methods, lateness becomes an issue of laboratory management practices, not sample
matrix.

If it is anticipated that some samples will have to be analyzed in a faster than normal turnaround
time during the performance of the analytical contract, a cost for these shorter turnaround time
samples should be negotiated prior to award of a contract. The bids should be broken out into
time periods that apply to the turnaround needs of the project (i.e., 2-day turnaround, 5-day
turnaround, 10-day turnaround, etc.).

3.4.3   Liquidated Damages and Penalties

In many cases, you should consider including penalty or damage clauses in your contracts as
incentives to preclude laboratories from defaulting on the contract, submitting data late, or
performing analyses improperly. Due to the nature of the services provided, it is often difficult to
assess actual damages caused by improperly performed analyses. Liquidated damages often are
used in many contracts in lieu of actual damages. Liquidated damages typically specify that, if the
laboratory fails to deliver the data specified in the deliverables section of the contract, or fails to
perform the services within the specified data turnaround time, the laboratory will pay a fixed,
agreed, price to compensate the organization  to whom the  services should have been delivered.
For example, some EPA contracts specify that the laboratory will pay, as fixed, agreed, and
liquidated damages, 2% of the analysis price per calendar day of delay, to a maximum reduction
of 50% of the analysis price.

If liquidated damages or penalties are involved, they should (1) be based on actual damage caused
(in  terms of cost) by each day of lateness, (2) be strong enough to discourage late delivery, and
(3)  be reasonable enough that they will not discourage laboratories from bidding. If liquidated
damages or penalties will be applied to meet the required data turnaround time, this information
should be included. The contract should specify that the laboratory will not be  charged with
liquidated damages when the delay in delivery or performance arises out of causes beyond the
control and without the fault or negligence of the laboratory. It also may be necessary to limit
damages to a certain dollar value or scope.

Other types of damages that should be considered and may be included in the contract include
costs for resampling, fines incurred as a result of improperly conducted analyses, and
administrative costs associated with the evaluation and processing of unacceptable data.

3.4.4  Reanalysis Costs

Every laboratory periodically produces data that are of little use for the intended purpose. While
well-run laboratories will contact the client as soon as they identify the problem and work with the
client to make the best of the situation, you still may find itself with no useful data and a deadline
approaching. The contract should stipulate that the laboratory will reanalyze samples at no cost to
the  client if the problems are due to laboratory error. It also should state that the client has the
right to inspect the results, and if they do not meet the requirements in the contract, the client has
the  right to reject the data, returning them to the laboratory without payment. Rejection of data
should be based on sound technical review of the results. It also obligates the client to make no
use of those results without making some payment to the laboratory.
                                            18

-------
                                                               Developing an Analytical Contract
3.4.5  Dilutions

The contract should discuss the instances in which dilutions of samples and reanalyses would be
considered billable by the purchaser. Again, a laboratory should be prepared to do the job right
the first time and not bill for reanalyses required due to their errors. In contrast, some samples
may need to be diluted and reanalyzed in order to bring the results within the demonstrated
calibration range of the instrumentation. This typically occurs when the concentration of
pollutants in the sample rums out to be higher than projected by the organization issuing the
contract. Dilutions also may be necessary when several pollutants are to be measured by a single
method, and the concentrations of some pollutants are within the calibration range of the
instrument but the concentrations of other pollutants are not. When this occurs, the laboratory
ought to be paid for their efforts to dilute the sample as necessary to quantify all pollutants. Such
reanalyses can be figured into the original price, inflating the per-sample price for all samples to
account for the need to reanalyze some samples, or it can be broken out as a separate cost. For
analyses involving an extraction or digestion as well as an analysis, it may be useful to specify the
price for the extraction step and the analysis separately, as it may be acceptable to simply dilute
and reanalyze the sample extract instead of diluting, re-extracting, and reanalyzing the entire
sample.

3.5   Developing a  Bid Sheet

After all project requirements have been established, you should develop a bid sheet to
accompany the  analytical requirements summary during the solicitation. The bid sheet allows
laboratories to submit bids in the same format, making bid evaluations easier, and also clarifies the
project. Bid sheets for analytical services typically are formatted as a chart, with analytical
requirements along one axis and number of samples and prices along the other. An example of a
bid sheet is attached as Appendix B.

The bid sheet should include the following information:

•      Project identifier

       Space for laboratory identification information

•      Day, date, and time of the bid deadline

•      Estimated award date
•      Laboratory period of performance (period of time during which the laboratory is obliged
       to resolve issues associated with analysis of the samples—generally six months after
       shipment of last sample)

•      Required delivery date (data turnaround time and the basis of its calculation, such as from
       receipt of each sample or from receipt of last sample)

       Bid validity period  (period of time during which bid prices are considered valid—generally
       45 days after the bid deadline; if the project is awarded after this period, you must contact
       bidding laboratories to determine if bids need to be revised)

       Parameters to be analyzed (typically the type of analysis and/or method)
•      Number of field samples to be analyzed for each parameter
                                            19

-------
Guidance on Procuring Analytical Services for Environmental Samples
       Number and type of billable QC samples (such as MS or SRM)

•      Total number of samples (field samples plus QC samples)

•      Columns for laboratories to submit per-analysis and total costs

Depending on the requirements of the project, additional information, such as qualification
information (discussed in detail in Section 4.3) may need to be requested with the bid sheet to
ensure that the laboratory will be able to meet the requirements of the analytical contract.

3.6   Estimating Costs

Before soliciting an analytical project, the anticipated cost of the work should be identified to
ensure that the solicitation  and procurement procedures are appropriate. Analytical projects
typically are costed-out using per-sample analysis prices. The most common methods for
estimating per-sample costs are: (1) reviewing current, published laboratory fee schedules for the
same or comparable analyses, and (2) reviewing historic per-sample costs for the same or
comparable analyses. Laboratory fee schedules are available by request from most commercial
laboratories. Previous invoice and payment records at your plant or organization can be used to
research historical costs. If your utility or company frequently outsources analytical work, it may
be helpful to copy the per-sample prices from these records into a separate file for future use in
estimating project costs and establishing the reasonableness of laboratory bid prices.
                                            20

-------
                                                            Soliciting and Awarding the Contract
       SOLICITING AND AWARDING THE CONTRACT
Procedures for soliciting and awarding contracts to perform analytical services can vary,
depending upon the scope of the project and purchasing requirements within the organization that
is issuing the contract. At one end of the spectrum are contracts that are awarded after placing a
single phone call and obtaining a quote from a single laboratory. The opposite end of the
spectrum are contracts awarded after a competitive solicitation and bidding process involving the
distribution of a detailed project description and a formal bid sheet via fax or mail. Determining
whether an analytical services request will be solicited on a casual basis, through a rigidly
documented formal solicitation, or somewhere in between, depends on the following factors:

       The nature of the analyses. Projects for routine analyses for which laboratories have
       published fee schedules are less problematic to solicit than projects for experimental or
       esoteric analyses. Phone solicitations to local laboratories or laboratories nationwide
       typically can be used for routine analyses to confirm laboratory prices. If the purchasing
       organization's procurement policies allow, an award can be made after per-sample prices
       are confirmed over the phone with a laboratory.

       The anticipated cost and the procurement system of the organization purchasing the
       analytical services. If the anticipated cost of the  project is minor and you do not have a
       highly structured procurement system, the most straightforward means of soliciting the
       project is to call one or more local laboratories, receive and evaluate the quotes, and
       award the work. However, if the anticipated cost of the project is substantial and/or the
       procurement system requires a competitive solicitation, enough laboratories should be
       solicited to  ensure that at least three bids are received (a minimum of three bids is required
       to qualify as a competitively awarded contract according to the Federal Acquisition
       Regulation  (FAR)). The project then can be awarded to the lowest of the three responsive,
       responsible bidders (Section 4.4).

•      Your knowledge of capable laboratories. If you frequently outsource projects to the
       same laboratory or laboratories, solicitations to these laboratories generally will not
       require the submission of prequalification data or references. Projects that are solicited to
       laboratories that are previously unknown to you may warrant additional steps, such as
       those described in Section 4.3, to ensure that the laboratory is capable of performing the
       requested analyses.

Because of the relatively straightforward nature of phone solicitations of routine projects, the
remainder of this chapter provides general guidelines for  conducting competitive, written
solicitations to multiple laboratories nationwide. Before implementing these procedures, you
should consult with their legal or procurement departments to ensure that the procedures are
consistent with those required within their organization.
                                           21

-------
Guidance on Procuring Analytical Services for Environmental Samples
4.1    Identifying Capable Laboratories and Transmitting the
       Requirements

Capable laboratories generally are defined as laboratories that have the instrumentation and
expertise to perform the analyses you require according to the methods you specify. Thus,
although a frequently used local laboratory may be perfectly capable of performing your routine
wet chemistry or metals work, that laboratory may not be considered capable when you require
samples to be analyzed for dioxins. Several laboratory indices are available as resources to enable
you to identify laboratories to target in a solicitation, including the ASTM International Directory
of Testing Labs, the American Council of Independent Laboratories' Directory, and the
DynCorp Directory of Environmental Testing Laboratories. Each of these directories is readily
available (see References 4 - 6).

After laboratories capable of performing the requested analyses are identified, a written bid
package needs to be transmitted to them. This bid package should include the analytical services
request and the bid sheet, at a minimum. The package also should include the required methods, if
non-routine analytical methods are required, and a cover letter if any additional or introductory
information needs to be provided to the laboratories. If possible, allow at least two weeks for the
laboratories to submit bids. This deadline should be noted on the bid sheet.

Traditionally, solicitation packages of 10 pages or less were transmitted by fax, and mail or
overnight services were used if the package was more than 10 pages. However, most laboratories
now have email addresses, and transmitting solicitations via email is  typically more efficient than
faxing or mailing the package.

4.2    Evaluating Bids

After the laboratories have received the solicitation and submitted their bids, you must evaluate
the bids to identify the laboratory that will be awarded the analytical services contract. Specific
procedures for evaluating bids may vary, depending upon the requirements of the organization
that is soliciting the contract. Therefore, it is recommended that the procedures used to evaluate
the bids be communicated  to all laboratories as part of the solicitation package.

One way to confirm the requirements will be met is to require the laboratories to submit a
technical proposal with their bids. An example technical proposal request and technical proposal
scoring sheet is provided in Appendix C. Generally, the bid evaluation process will focus on an
evaluation and comparison of each laboratory's  proposed cost and an evaluation of each
laboratory's capability to meet the analysis requirements.

You  should consult your legal department or purchasing department to identify any applicable
requirements for evaluating competitive bids within their organization. In the absence of explicitly
defined bid evaluation procedures, you may wish to follow the procedures outlined below. These
procedures, which have been adapted from those published in the FAR, begin with evaluation of
all bids received to identify the lowest responsive, responsible bid. A bid is considered responsive
if the following criteria are met:

•       The bid was submitted without contingencies or with acceptable contingencies
                                            22

-------
                                                             Soliciting and Awarding the Contract
       The bid was submitted before the bid deadline

       The bid sheet (if required) contains no errors or omissions

The organization responsible for awarding the contract also should recalculate bid prices based on
each laboratory's per-sample price to ensure that the bidding laboratories did not make any
mathematical errors. If any incorrect calculations are identified, the laboratory  should be
contacted to confirm the corrected total bid price. In addition, you should ensure that there are no
unacceptable contingencies associated with any of the bids (such as the use of an unacceptable
method). After all bids have been checked for errors and contingencies, you can identify the
lowest, responsive bidding laboratory for the project. If there is a question regarding a
laboratory's ability to perform the work, you should perform a responsibility determination, as
well (see Section 4.3).

If three or more responsive bids were received, then the low bid may be deemed reasonable based
on the closeness of the bid prices to each other and current market conditions. If fewer than three
bids were received, price reasonableness can be determined using bid prices submitted for
comparable projects, price quotes from current laboratory fee schedules, or information requested
from the laboratory, including a breakdown of costs or invoices to other clients for comparable
work. The lower bid may be deemed unreasonable if it is significantly lower than the other bids,
and may not be considered for award because it can indicate a lack of understanding of the
requirements.

4.3   Conducting Responsibility Determinations

If the low-bidding laboratory is previously unknown to you, the importance of the project merits
special effort to ensure that the awarded laboratory is capable of reliably performing the requested
analyses, or you are required to follow the FAR, then a laboratory responsibility determination
should be performed. The best means of confirming that a laboratory is capable of reliably
performing an analytical requirement is to assess data recently produced by the laboratory using
the same method on similar sample matrices. A  less expensive approach is to rely on other
information applicable to the analyses in question, such as performance evaluation (PE) sample
results, federal or state certifications, and corporate references. Sections 4.3.1  through 4.3.4
provide guidance for using the laboratory's method performance data, PE sample results,
certifications, or references to evaluate their capability.

If laboratory performance cannot be  assessed based on existing data or references, another
alternative is to require laboratories that bid on the project to analyze samples specific to the
project and submit these results with their bids. Bids then are evaluated in terms of cost and
performance. Laboratories that do not submit acceptable data are not qualified to perform work
under the project, and can be eliminated from consideration for award. Section 4.3.5 provides
additional guidance concerning the use of prequalification analyses as a means  of evaluating
laboratory capability.

For long-term, critical, or very costly projects, you should consider auditing the laboratory before
an award is made. Section  4.3.6 provides guidance on conducting audits.

4.3.1  Method Performance Data
                                            23

-------
 Guidance on Procuring Analytical Services for Environmental Samples
Many laboratories routinely use EPA-approved methods for analysis of wastewater and drinking
water samples collected by their clients (including the methods approved at 40 CFR Parts 136,
141, and 405 - 471). In such cases, you can ask a laboratory to provide historical data that
demonstrates the laboratory is capable of reliably analyzing the required sample matrices with the
required methodology. Data requested should include results from all QC parameters required by
the method, including results from calibration standards, blanks, initial and ongoing precision and
recovery samples, and spiked matrix samples. You should request historical data generated within
the past six months. Older data still may be relevant, but the laboratory should indicate any
personnel, instrument, or facility changes that have occurred since the data were generated.

4.3.2  Performance Evaluation Sample Results

Several EPA and state laboratory programs send performance evaluation (PE) samples to
laboratories that are part of their program on a periodic or regular basis to monitor laboratory
performance. PE samples typically consist of a synthetic matrix spiked with concentrations of
analytes known to the program office but unknown to the laboratory (single-blind samples). The
program laboratories analyze the samples and report the results, and the program office compares
these results to the true values of the PE samples. The program office or  laboratories that
participate in programs that issue PE samples should be able to provide you with the assessment
of their latest PE sample results.

Several PE studies programs are administered by EPA in support of the Clean Water Act, the
Safe Drinking Water Act, and Superfund:

•      Water Pollution (WP). Laboratories in the WP program  receive  chemistry PE samples;
       the program tests laboratories'  abilities to analyze for common surface water quality
       parameters and pollutants. The WP program supports more than 25 state wastewater and
       other environmental laboratory certification programs.

•      Discharge Monitoring Report Quality Assurance (DMRQA). Laboratories in the
       DMRQA program receive chemistry and whole effluent toxicity PE samples. This national
       program is used by  EPA and the states to ensure the quality of monitoring data submitted
       by more than 7,000 major NPDES permittees each year.

•      Water Supply (WS). The Water Supply program includes chemistry, microbiology, and
       radiochemistry PE studies and  supports  the Safe Drinking Water Act.

•      Effluent Guidelines Program. Laboratories awarded contracts to analyze samples for
       EPA's Engineering Analysis Division within the Office of Water's Office of Science and
       Technology are sent periodic PE samples for organics, metals, and wet chemistry analyses
       to monitor performance.

•      Contract Laboratory Program (CLP). Laboratories in EPA's Contract Laboratory
       Program, which supports Superfund sample analyses, receive PE samples for organics and
       inorganics analyses  on a quarterly basis.

Other PE programs administered by private associations also may be helpful in meeting your
needs. The European Union, for example, has developed the QUASIMEME program to provide a
means for private organizations interested in participating in laboratory performance studies. The
program offers a wide variety of PE samples and intercomparison studies on a routine basis.

                                           24

-------
                                                             Soliciting and Awarding the Contract
Information on this program is available by email at marlab.ac.uk, by phone at 044 (0) 1224295
352, and by fax at 0044 (0) 1224 295511.
In general, a laboratory not participating in a PE sample program or equivalent should not be
considered for the contract. However, it is important to understand that PE sample results are
only useful if the analyses are applicable to the project for which the laboratory is considered. A
laboratory's ability to perform well on organics PE samples is not an indication of how reliable its
metals laboratory is.

4.3.3  Certifications

You also can ask laboratories to supply a list of their current certifications, such as state
certifications. In addition, information about laboratory certifications can be obtained through
Internet searches or by telephone or email from the NPDES or drinking water staff or QA officer
in the state or EPA regional office with jurisdiction over the certified laboratory.

State certification programs vary widely—some states certify for specific programs, such as
drinking water monitoring, while others certify for specific methods or analyses. Certifications are
particularly useful  if they apply directly to the analyses you required, but also provide an
indication of the overall standing of the laboratory. Most certification programs entail laboratory
audits and PE sample analyses, and thus provide some assurance that the laboratory is generally
capable of providing  reliable analytical services. However, you should note that a state drinking
water certification  is  no guarantee that a laboratory is capable of performing industrial wastewater
analyses by methods  not covered by that certification. Conversely, the absence of a certification is
no guarantee that the laboratory is not capable of performing the analyses—a certification for that
method or parameter simply may not exist.

Currently, guidance and standards for a national laboratory accreditation program are being
developed through a  state/EPA organized  group known as the National Environmental
Laboratory Accreditation Program (NELAP). Current information on NELAP and the National
Environmental Laboratory Accreditation Conference (NELAC) is available on the Internet.

4.3.4  References

This means of establishing a laboratory's reliability and capability is, perhaps, the easiest. If you
have not worked with a particular laboratory before, you can ask the laboratory to provide
contacts and phone numbers of corporate or government clients for which the laboratory has
performed services comparable to the project at hand. Questions to ask the references include:

•       Did the laboratory provide data by the required due date?

•       Were the data reviewed upon receipt to ensure that the laboratory performed the
       requested analyses according to the specified methods and with the required QA/QC ? (If
       the answer to this  question is no, the reference is not likely to be capable of providing
       sufficient information to adequately assess the laboratory's capability.)

•       Does the laboratory have a documentation system for sample control that retains accurate
       records of chain-of-custody; sample holding, handling, preservation, and analyses; raw
       data; QA/QC, and processed data? Have you audited this system?
                                            25

-------
Guidance on Procuring Analytical Services for Environmental Samples
•      Were laboratory personnel easy to work with when problems arose during all phases of
       the project, including sample scheduling, sample analysis, and data review? If problems
       were noted during data review, was the laboratory prompt and responsive in addressing
       your concerns?

       Do you have any reservations in recommending this laboratory?

4.3.5  Prequalification Analyses

As noted above, prequalification analyses may be required if laboratory performance cannot be
assessed based on existing data, certifications, or references. Two options are available regarding
payment of prequalification analyses. The first is to require laboratories to provide prequalification
data at no cost with their bids. Laboratories can recoup this cost if they are awarded the contract.
This approach generally will not work if the project is small, and the laboratory has little incentive
to provide prequalification data at no cost. If, however, the project entails analysis of a sufficient
number of samples to justify a loss leader from the laboratory, this approach should be
considered.

The second option entails payment for prequalification analyses. In such  a situation, laboratories
would bid on the project in two parts: one portion of the bid would apply only to prequalification
analyses, while the balance of the bid would apply only to analysis of the real samples. The bids
would be evaluated based on overall cost, and the laboratories with the lowest cost would be
awarded contracts to perform only the prequalification analyses. After prequalification data have
been submitted and evaluated, the lowest bidding laboratory with acceptable prequalification data
would be awarded the contract to analyze the real samples during the balance of the project.

Prequalification analyses can take several forms, including:

       Analysis of single-blind samples.  The best way of determining laboratory performance
       before award is requiring bidding laboratories to analyze samples that are spiked with the
       target analyte(s) at concentrations unknown to the bidding laboratories. Such samples are
       essentially identical in concept to the PE samples described in 4.3.2. You can either
       prepare their own  single blind samples or you can purchase these  samples from
       commercial vendors.

       Vendors typically carry several types of stock PE samples applicable to a variety of
       pollutants, matrices, and analytical methods. To ensure that laboratories are unable to
       "predict" the pollutants and associated concentrations in their PE samples, vendors offer
       PE samples that contain a minimum number of pollutants from a selected list (such as at
       least 7 of 10 listed metals), each of which will be present within a specified "range" (such
       as 1 - 50 Mg/L)- Vendors routinely prepare and distribute new batches in order to further
       protect the integrity of their PE sample program. Actual pollutants and pollutant
       concentrations in each batch are certified, and these "certified values" are provided to the
       organization that purchases and distributes the PE sample(s). You should purchase the PE
       sample that most closely matches their target pollutant list and concentration range.

       Analysis of samples spiked at the laboratory. A simpler, and potentially more cost-
       effective approach to the single-blind sample analysis  scenario is to require laboratories to
                                            26

-------
                                                             Soliciting and Awarding the Contract
       spike samples in-house and provide the spiking levels and recoveries for evaluation. If this
       approaches chosen, it is recommended that the laboratory be required to spike and analyze
       four replicate samples so that both precision and accuracy can be assessed. The matrix
       used can include reagent water, effluent provided by the wastewater plant or source or
       finished water from the drinking water plant, or a representative matrix that can be
       selected by the laboratory. If the laboratory is permitted to select a matrix type for this
       analysis (such as municipal wastewater or ambient water), the data reported should
       include characterization data, such as turbidity, hardness, background concentrations of
       the unspiked sample, etc.

•      Analysis of a standard reference matrix. A third, similar approach is analysis of a
       commercially available SRM. The SRMs should be chosen by the client, and can be
       purchased by the laboratory or purchased by the client and sent to the laboratory for
       analysis.

•      Analysis of method blanks and method detection limit studies. If the project entails
       detection of analytes at very  low levels, the laboratory(ies) awarded the project should be
       required to demonstrate that  laboratory contamination does not exceed acceptable levels
       and demonstrate that they are capable reaching the low end of the detection range. The
       latter is accomplished by performing a method detection limit study according to the
       procedure at 40 CFR Part 136 Appendix B (essentially, analysis of seven replicate reagent
       water samples spiked with the analyte of interest at one to five times the method's
       estimated detection limit). A  method blank analyzed with these MDL samples can be used
       to demonstrate freedom from contamination at low levels.

4.3.6  Laboratory Audits

The goal of a prequalification audit is to ensure that the laboratory has the capability and
commitment to meet the program goals of timely delivery and high-quality analytical services.
Audits may be announced, or an alternate technique to determining that a laboratory is capable of
reliably performing the contract is to make an unannounced visit to the laboratory. Audits can
focus on  any or all of the following areas:

•      Laboratory personnel qualifications
       Sample receiving and storage areas

•      Sample preparation and analysis areas

•      Instrumentation

       Laboratory quality assurance plan (QAP)

•      Laboratory standard operating procedures (SOPs)

Although laboratory audits generally are specific to the project, general criteria are applicable to
each of the above areas. The best approach to evaluating a laboratory, based on these criteria, is
through the use of checklists. Examples of laboratory audit checklists are provided in Appendix
D. These checklists should be modified as necessary to adapt them to the specific project.
                                            27

-------
Guidance on Procuring Analytical Services for Environmental Samples
Contact the appropriate state or EPA regional office via phone or email to obtain audit or
inspection information about the laboratory. The Internet can be used to identify the appropriate
state or EPA regional contact.

You have two options if a laboratory fails an audit: (1) the laboratory can be eliminated from
consideration for the project, or (2) the laboratory can be provided the opportunity to correct the
deficiencies identified in the audit and request a reevaluation. If the laboratory passes the audit,
you can proceed to contract award.

4.4    Awarding the  Contract

Contract awards typically should be made over the phone, then followed by a written contract for
laboratory signature. Awarding the  contract over the phone enables you to verify the scope of the
analytical work and verify laboratory information. This information should include the name of the
person assigned to receive the samples and the street address to which the samples will be
shipped—overnight  delivery services, such as Federal Express, will not accept samples with post
office box addresses. Laboratory information also should include the name and address of the
laboratory's administrative personnel that handle billing issues, as these may differ from the
address to which samples are shipped.
                                            28

-------
                                       Transporting Samples and Communicating with the Laboratory
5     TRANSPORTING SAMPLES AND COMMUNICATING WITH THE
       LABORATORY
After the analytical services contract is awarded, samples are collected and shipped to the
laboratory. Although it is the laboratory's responsibility to contact the client if problems occur
after sample receipt, you still should initiate communications with the laboratory periodically to
monitor progress.

5.1    Transporting and Tracking Samples

You must ensure sample integrity from collection through data reporting to use the data for
anything other than internal purposes. This includes the ability to trace possession and handling of
the sample from the time of collection through analysis. The following items and steps will ensure
that samples are processed accurately and that the data produced are defensible: sample labels,
sample seals, field log books, chain-of-custody records, sample analysis request sheets, tracking of
sample delivery to laboratory, receipt and logging of samples by the laboratory, and
documentation of the sampling  project from sample collection through sample analysis. This
process of tracking samples is considered a "sample control system," and should be established as
a documentation system for the laboratory.

•      Sample labels. Sample labels always should be used to prevent sample misidentification.
       The sample number and required analyses should be stated clearly on the label. If space
       allows, the name of sampler, date and time of collection, and place of collection also
       should be included. Waterproof markers should be used to write on sample labels.

•      Sample seals. When chain-of-custody is critical,  sample seals can be used to detect any
       unauthorized tampering with samples up to the time of analysis. The seal should be
       attached in such a way that it is necessary to break it to open the sample container.

•      Field log book. A field log book should be used  to record all information pertinent to
       sample collection. The field log book should include the following:  the purpose of
       sampling, the location of the sampling point, the name and address of the field contact, the
       producer of the material being sampled and address (if different from sampling location),
       the type of sample being collected (such as wastewater, drinking water, or biosolids), and,
       if the sample is a wastewater, the identification of the process producing the waste stream.
       In addition, the number of samples and volume of sample taken, the description of the
       sampling point and sampling method, the date and time of collection, and the sampling
       label number should be  included. Other items that are useful to keep with the field log
       book are references such as maps or photographs of the sampling site, field observations
       and measurements, and  signatures of personnel responsible for observations. Sampling
       situations vary, so no general rule can be given as to the information to be entered in the
       log book, but as much information as possible should be provided.

•      Chain-of-custody record. The ability to trace possession and handling of a sample from
       the time of collection through analysis is referred  to as chain-of-custody. A sample is

                                          29

-------
Guidance on Procuring Analytical Services for Environmental Samples
       considered to be in an individual's custody if any of the following criteria are met: (1) the
       sample is in your possession or it is in your view after being in your possession, (2) it was
       in your possession and then locked up or sealed to prevent tampering, or (3) it is in a
       secured area. The chain-of-custody record is used  as physical evidence of sample custody.
       The sampler completes a chain-of-custody record  to accompany each sample or group of
       samples shipped from the field to the laboratory. The record includes the following:
       sample number, signature of sampler, date, time, and location of collection, sample type,
       signatures of persons involved in the chain of possession and inclusive dates of possession.
       The original signature copy of the chain-of-custody record is enclosed in plastic and
       secured to the inside of the container used for sample shipment. A copy  of the custody
       record is retained for the sampler's  file. The shipping containers are secured and custody
       seals are placed across the cooler openings. The laboratory representative who accepts the
       incoming sample shipment signs and dates the chain-of-custody record to acknowledge
       receipt of the samples. Custody  procedures continue to be followed throughout the
       laboratory (from sample custodian to extracting specialist to analyst) until the sample is
       consumed or disposed of.

       Sample analysis request sheet. A sample analysis request sheet or traffic report should
       accompany the samples to the laboratory. The  sampler should complete  the field portion
       of this sheet with most of the pertinent information noted in the log book. The laboratory
       representative should complete the laboratory portion of this form, which includes: the
       name of the person receiving the sample, laboratory sample number, date of sample
       receipt, condition of samples upon receipt, and analyses to be performed.

       Sample delivery to laboratory. The samples  should be delivered to the laboratory as
       soon as practicable. Commercial carriers often are the best method of shipment if the
       samples cannot be delivered to the laboratory the same day as collection. It also is
       advisable to use a carrier with the ability to track and provide the status of individual
       shipments. To facilitate return of the shipping containers, shippers should clearly mark the
       name and address of the return destination on the containers. The laboratory must be
       contacted every day they are to receive samples to confirm receipt of samples. Both you
       and the laboratory should document this confirmation.

       Receipt and log-in of samples. At the laboratory, the sample custodian receives the
       samples and should perform the  following tasks with each sample: (1) inspect the
       condition of the sample, (2) inspect the condition of the sample seal (if present), (3)
       reconcile sample label information and seal against the chain-of-custody  record, (4) assign
       a laboratory sample number, (5) log the sample in the laboratory log book, and (6) store
       the sample in appropriate storage conditions.

       Documentation of sampling project from sample collection through sample analysis.
       Documentation of the entire sampling project from sample collection through sample
       analysis, including any problems and resolutions that occur during the event, should be
       maintained.

       Sample holding times. Sample  analysis results may not be valid if the prescribed holding
       times and other requirements for each parameter are not met. These requirements are
       listed in Table II of 40 CFR Part 136 for wastewater contaminants, as well as in Subpart C
       of 40 CFR Part 141  for many drinking water contaminants.
                                           30

-------
                                        Transporting Samples and Communicating with the Laboratory
Samples should be packaged for shipment in compliance with the most current U.S. Department
of Transportation, state, local, and commercial carrier regulations. All required government and
commercial carrier shipping papers must be completed and shipment classifications made
according to these regulations.

Waterproof, metal or hard plastic ice chests or coolers should be used for shipment. Inside the
cooler, sample containers should be enclosed in clear plastic bags so that sample tags and labels
are visible. Water and soil samples suspected to contain dioxin or highly  toxic or reactive
pollutants at high concentrations must be enclosed in a metal can with a clipped or sealed lid
(paint cans typically are used). The outer metal can must be labeled with  the number of the sample
contained inside. Containers that do not fit into paint cans should be double bagged.

Ideally, shipping containers  should be packed with noncombustible, absorbent packing material,
such as vermiculite. The material should surround the sample bottles or metals cans containing
sample to prevent breakage  during transport. Earth or loose ice should never be used to pack
samples; earth is a contaminant, and ice melts, resulting in container breakage.

The sampling and shipping conditions for each sample will depend on the analysis required for
that sample, and will be specified in the method. When shipping with ice, the ice should be  in
sealed plastic bags to prevent melting ice from soaking packing material which, when soaked,
makes  handling of samples difficult in the laboratory. The Sample Analysis Request Sheet,  chain-
of-custody record and any other sample documentation accompanying the shipment must be
enclosed in a waterproof plastic bag and taped to the underside of the cooler lid. Coolers should
be sealed with custody seals in such a manner that the custody  seal would be broken if the  cooler
were open. Shipping coolers must have clearly visible return address labels on the outside.

Samples should be shipped through a reliable commercial carrier, such as Federal Express,  Emery,
and Airborne Express, or equivalent if the samples cannot be delivered to the laboratory by the
sampler on the day or day after the sampling occurs. The sampler should record the shipment
tracking number (such as the airbill number) to resolve any problems associated with the  sample if
it is waylaid. Consideration also should be given to requesting the laboratory to pick up the
samples.

5.2   Communicating with the Laboratory

You must maintain communications with the laboratory to confirm sample shipment receipt,
timely  analysis, and quality data. In addition, it is important that the laboratory is able to
communicate immediately with the sampler or person responsible for the  sampling event  in case of
sample shipment problems or analysis issues that may affect data quality.

Although phone communications currently are the norm, these communications ideally should be
conducted via email. Email communications not only should provide virtually immediate
responses, but also enables both the contracting party and the laboratory  to maintain a written
record of sample receipt confirmations, problem notifications, and problem resolutions. In
addition, email communications reduce misunderstandings and miscommunications.
                                           31

-------
                                                                   Reviewing Analytical Data
       REVIEWING ANALYTICAL DATA
When reviewing data submitted by contract laboratories, you must ensure the test data include the
QA/QC elements listed in the analytical method and in your contract (see Section 3.3);otherwise,
the data can be considered noncompliant. As a result,. These supporting QA/QC results provide
you with the simplest means of assessing the quality of your data.

In many of its early analytical programs, EPA relied upon laboratories to maintain records of the
QA/QC data. This practice was cumbersome for the laboratories, because many of the QA/QC
data were common to the analytical results for a variety of clients. Retrieving these data from the
laboratory to resolve questions of permit compliance was time-consuming for the permittee and
the permit writer. More importantly, this practice occasionally resulted in unscrupulous
laboratories  failing to perform the necessary QA/QC testing, or performing the QA/QC testing
"after the fact" to satisfy an audit or data submission request. In particular, many laboratories did
not perform the IPR test prior to practice of the method and did not perform a spike of the
analytes into the sample matrix to prove that the method would work on a particular sample.
Therefore, while the data provided by those laboratories may have been valid, there was no way
to prove their validity.

Sections 6.1 through 6.11, below, provide guidance on evaluating sample data based on QA/QC
data. A data inspection checklist is provided in Appendix E, providing a standardized format for
the data review process and the documentation of findings.

6.1   Purity and Traceability of Reference Standards

Laboratories submitting analytical data must be able to trace the reference standards used in the
analysis to EPA or NIST. The proof of this traceability is a written certification from the supplier
of the standard. Documentation of the purity and traceability of the standards need not be
provided with every sample analysis. Rather, it should be maintained on file at the laboratory and
provided on request. When analyses are conducted in a contract laboratory,  such documentation
ought to be provided to the permittee the first time that a laboratory is employed for specific
analyses and then updated as needed.

6.2   Calibration Range

The data reviewer must make certain that the calibration range encompasses the minimum level
and that all measurements are within the calibration range of the instrument. Samples with
analytes outside of the calibration range should be diluted and reanalyzed. The diluted sample
results need only apply to those analytes that exceeded the calibration range  in the initial analysis.
In other words, it is acceptable to use data for different analytes from different levels of dilution
within the same sample.
                                           33

-------
 Guidance on Procuring Analytical Services for Environmental Samples
 If data from an analysis of the diluted sample are not provided, limited use can be made of the
 data that are above the calibration range. The response of the analytical instrument to
 concentrations of analytes will eventually level off at concentrations above the calibration range.
 While it is not possible to specify at what concentration this will occur from the calibration data
 provided, it is generally safe to assume that the reported concentration above the calibrated range
 is a lower limit of the actual concentration. Therefore, if concentration above the calibration range
 is also above a regulatory limit, it is highly likely that the actual concentration would also be
 above that limit.

 6.3    Linearity of Calibration

 Linearity specifications vary from method to method, depending on the  quantitation technique.
 Typical limits on the RSD are as follows:

        15% for GC and HPLC methods

       35% for analytes determined by the internal standard technique in GC/MS methods

 •      20% for analytes determined by isotope dilution in GC/MS methods

 If the calibration is not linear, as determined by the RSD of the response factor or calibration
 factor, the calibration curve, as opposed to the average response factor,  must be used for
 quantitation. This means that a regression line or other mathematical function must be employed
 to relate the instrument response to the concentration. Properly maintained  and operated lab
 instrumentation should have no difficulty in meeting linearity specifications for the EPA-approved
 methods.

 Whatever calibration range  is used, the laboratory must provide the RSD results by which one can
judge linearity, even in instances where the laboratory is using a calibration  curve. In instances
 where the laboratory employs a curve rather than an average response factor, the data reviewer
 should review each calibration point to assure that the response increases as the concentration
 increases. If it does not, the instrument is not operating properly, or the calibration  curve is out of
 the range of that instrument, and data are not considered valid.. The analysis of samples should
 not proceed until linearity on that instrumentation is demonstrated.

 6.4    Calibration Verification

 Calibration verification results should be within method specifications. If any individual value falls
 outside the range given, system performance is considered unacceptable, and the laboratory may
 either recalibrate the instrument or prepare a new calibration standard and make a second attempt
 to verify calibration. If the laboratory was not able to verify calibration, the  data should be
 evaluated to determine if it is usable with a qualification of high or low bias, or if the bias
 precludes use of the data.

 6.5    Method Detection Limit,  Minimum Level, or Quantitation Limit

 Unless specific data gathering requirements require otherwise, the laboratory should report the
 concentration of all sample  results that are at or above the ML. It should be noted that this ML is
 a sample-specific ML and, therefore, reflects any sample dilutions that were performed. If sample
                                            34

-------
                                                                    Reviewing Analytical Data
results are reported below the ML, the data reviewer should require the responsible party to
correct and resubmit the data, or if this course of action is not possible, the reviewer should
determine the sample-specific ML and consider results below that level to be non-detects for
regulatory purposes.

If sample results are reported above the ML, but are below a compliance level, then the data
reviewer should consider the results to suggest that the pollutant has been detected but is
compliant with the facility's permit (assuming that all QC criteria are met). If sample results are
reported above a compliance level, the data reviewer must evaluate laboratory QC samples in
order to verify that the level of pollutant is not attributable to analytical bias. In addition, the data
reviewer must evaluate all blank sample results in order to determine if the level of pollutant
detected may be attributable to contamination.

Although sample results are to be reported only if they exceed the ML, all blank results are to be
reported, regardless  of the level. This reporting requirement allows data reviewers the opportunity
to assess the impact  of any blank contamination on sample results that are reported above the ML.
6.6    Initial Precision and Recovery

If the IPR data fail to meet the specifications in the method, none of the data produced by the
laboratory can be considered to be valid. If the laboratory did not perform the start-up tests, the
data cannot be valid, unless all other QC criteria have been met and the laboratory has submitted
IPR (and associated instrument QC) data that were generated after-the-fact by the same analyst
on the same instrument. If these conditions are met, then the data reviewer may consider the data
to be acceptable for most purposes. NOTE: The inclusion of this alternative should not in any
way be construed to sanction the practice of performing IPR analyses after the analysis of field
samples. Rather, EPA believes that demonstration of laboratory capability prior to sample analysis
is an essential QC component; this alternative is provided only as a tool to permitting authorities
when data have already been collected without the required IPR samples. Once the problem has
been identified, all responsible parties are expected to implement corrective action necessary to
ensure that it is not repeated.

It is important to remember that if a change is made to a method, the IPR procedure must be
repeated using the modified procedure. If the start-up test is not repeated when these steps are
modified or added, any data produced by the modified methods cannot be considered to be valid.
Such changes may involve alternative extraction, concentration, or cleanup processes; alternative
GC columns, GC conditions, or detectors; or other steps designed to address a particular matrix
problem..

6.7   Ongoing  Precision and  Recovery

The data reviewer must verify that the OPR sample has been run with each sample batch and that
the  applicable recovery criteria in the analytical method have been met. If the recovery criteria
have not been met, the reviewer may use the following guidelines when making use of the data:
                                           35

-------
Guidance on Procuring Analytical Set-vices for Environmental Samples
•      If the concentration of the OPR is above method specifications but that analyte is not
       detected in an associated sample, then it unlikely that the sample result is affected by the
       failure in the OPR.

       If the concentration of the OPR is above method specifications and that analyte is detected
       in the sample, then the numerical sample result may represent an upper limit of the true
       concentration, and data users should be cautioned when using the data for enforcement
       purposes.

       If the concentration of the OPR is below method specification but that analyte is detected
       in an associated sample, then the sample result may represent the lower limit of the true
       concentration for that analyte.

•      If the concentration of the OPR is below method specification and that analyte is not
       detected in an associated sample, then the sample data are suspect and cannot be
       considered valid for regulatory compliance purposes.

If the OPR standard has not been run, there is no way to verify that the laboratory processes were
in control. In such cases, a data reviewer may be able to utilize the field sample data by examining
the matrix spike recovery results (see item 9), the IPR results, OPR results from previous and
subsequent batches, and any available historical data from both the laboratory and the sample site.
If the matrix spike results associated with the sample batch do not meet the performance criteria
in the methods, then the results for that set of samples cannot be considered valid. If the
laboratory's IPR results and the matrix spike results associated with the sample batch in question
meet the all applicable performance criteria in the methods, then the data reviewer may be
reasonably confident  that laboratory performance was in control during field sample analysis. This
level of confidence may be further increased if there is a strong history of both laboratory
performance with the method and method performance with the sample matrix in question, as
indicated by additional OPR and matrix spike data collected from the laboratory and samples from
the same site.
6.8   Analysis of Blanks

Unless the samples are still within analytical holding time and reanalysis is possible, there is no
corrective action if unacceptable blank data are submitted with sample data. Therefore, the
reviewer has several options in making use of the sample data. First, if a contaminant is present in
a blank, but not present in a sample, then there is little need for concern about the sample result,
though it may be useful to occasionally review the raw data for samples without the contaminant
to ensure that the laboratory did not edit the results for this compound.

The second approach deals with instances where the blank contaminant is also reported in a
sample. Some general guidance will help you determine the degree to which the contaminant is
affecting sample results:

       If the sample contains the contaminant at levels of at least 10 times that in the blank, then
       the likely contribution to the sample from the contaminant in the laboratory environment is
       at most 10%. Since most of the methods in question are no more accurate than that level,
       the possible contamination is negligible.
                                            36

-------
                                                                    Reviewing Analytical Data
 •      If the sample contains the contaminant at levels of at least 5 times but less than 10 times
       the blank result, the compound is probably present in the sample, but the numerical result
       should be considered an upper limit of the true concentration.

 •      If the sample contains the contaminant at levels below 5 times the level in the blank, there
       is no adequate means by which to judge whether or not the sample result is attributable to
       laboratory contamination. The results for that compound in that sample are then suspect.

 There are two difficulties in evaluating sample results relative to blank contamination. First, the
 reviewer must be able to associate the samples with the correct blanks. The second difficulty
 involves samples that have been diluted. The dilution of the sample with reagent water or the
 dilution of the extract with solvent represents an additional potential source of contamination that
 will not be reflected in the results for the blank unless the blank was similarly diluted. Therefore,
 in applying the 10-times rule, the concentration of the sample is compared to the blank result
 multiplied by the dilution factor of the sample or sample  extract. For instance,  if 12 ppb of a
 contaminant are found in the blank, and the associated sample extract was diluted by a factor of 6
 relative to the extract from the blank prior to analysis, then the sample result would have to be
 greater than 12x6x10, or 720 ppb, to be acceptable. Between 360 ppb and 720 ppb, the sample
 result would best be considered an upper limit of the actual concentration. Below 360 ppb, the
 sample result is not acceptable for compliance monitoring.

 In most cases, the practice of subtracting the concentration reported in the blank from the
 concentration in the sample is not recommended as a tool to evaluate sample results associated
 with blank data. One of the most common problems with this approach is that  blank
 concentrations are sometimes higher than one or more associated sample results, yielding negative
 results.

 6.9   Matrix Spikes and Labeled Compound Spikes

 When evaluating matrix spike results, the data reviewer must verify the following:

 •      An appropriate spike concentration was used

 •      The unspiked sample has been analyzed

       The spiked sample has been analyzed

 •      The recovery of the spike is within the range specified

 •      If the spike recovery is not within the range specified, a QC check standard has been
       analyzed
 •      If a QC  check standard has been analyzed, the results are within the range specified

For isotope dilution analyses, the evaluation of the data is simpler because isotopically labeled
analogs of the pollutants are spiked into each sample allowing recovery to be evaluated for every
analyte in every sample, and because a QC check standard (termed the "ongoing precision and
recovery standard," or OPR) is analyzed with each sample set.

If the recovery of a labeled compound spiked into the sample is not within the range specified in
the method, and the results of analysis of the ongoing precision and recovery standard are within
                                           37

-------
Guidance on Procuring Analytical Services for Environmental Samples
the respective limits, the sample results are considered reportable, with qualification that the
results may be biased. When labeled compound recoveries are outside of the method
specifications, the problem may be related to the sample matrix. In these instances, the sample
may be diluted with reagent water and reanalyzed. If the labeled compound recoveries meet the
method specifications after dilution of the sample, then the results are acceptable, although the
sensitivity of the analysis will be decreased by the dilution.

In instances where matrix spike or labeled compound recoveries are not within the specifications,
it may still be possible to use the sample results for compliance monitoring purposes. In particular,
if (1) the recovery of the spiked compound is above the method specifications and (2) the
compound is not detected in the sample analysis, it is unlikely that the compound is present in the
sample. This is because the factors that caused the analysis to over-estimate the concentration in
the spiked sample would not likely have resulted in an under-estimate in the unspiked sample. For
samples in which the compound is detected but the matrix spike or labeled compound recovery is
above the method specifications, the concentration reported in the unspiked sample is likely an
upper limit of the true concentration.

Unfortunately, for some sample matrices, even dilution will not resolve the problem, and for other
matrices, the loss of sensitivity will preclude the use of the results for determining compliance. In
these instances, additional steps need to be taken to achieve acceptable results.

6.10  Statements of Data Quality for Recovery of Spiked Analytes or
       Labeled Compounds in Samples

Many laboratories do not provide the data quality statements with the sample results, in which
case the data reviewer must determine if the data quality statements are being maintained for each
analyte and may need to obtain the data. If necessary, the reviewer can construct the data quality
statement from the individual data points.

The lack of a statement of data quality does not invalidate results but makes some compliance
decisions more difficult. If statements of data quality are not being maintained by the laboratory,
there may be increased concern about both specific sample results and the laboratory's overall
quality assurance program.

6.11  Field Duplicates

The field duplicate provides an indication of the overall precision associated with entire data
gathering effort, including sample collection, preservation, transportation, storage, and analysis
procedures. The data reviewer should examine field duplicate results and use the following
equation  to calculate the relative percent difference between the duplicate and its associated
samples.
                                              (D1+D2)
                                           38

-------
                                                                      Reviewing Analytical Data
       where:
                Dl = concentration of the analyte in the field sample
                D2 = concentration of the analyte in the duplicate field sample

If the analyte of interest was not detected in either replicate of the field sample, then the RPD will
be zero. If the analyte was detected in each field sample replicate, but the results are highly
disparate (indicated by a large RSD),  the reviewer should apply the following guidelines when
making use of the data:
       If the analyte was detected in  each replicate and at similarly variable concentrations in the
       blank samples, then the field sample variability may be attributable to variable
       contamination, and the data may not be valid for regulatory compliance purposes.

       If the analyte was detected in each replicate at a concentration well above the regulatory
       compliance level, but was not detected in the associated blank samples, then it is likely that
       the sample results are not adversely affected.

Ideally, the RPD between field duplicates and MS/MSD samples will be close to zero. Any
difference between the two duplicates is  attributable to variability associated with the field
sampling process.
                                            39

-------
      	References

       REFERENCES
1      U.S. EPA Office of Water. Environmental Monitoring Methods Index. Version 2.0.
          National Technical Information Service (PB97-502637BDR), 1995

2      U.S. EPA Office of Water. Methods and Guidance for Analysis of Water. EPA 821 -C-
          978-00 I.April 1997.

3      International Organization for Standardization (ISO) and International Electrotechnical
          Commission (IEC). 1990. Guide 25: General Requirements for the Competence of
          Calibration and Testing Laboratories.

4      American Society for Testing and Materials. 1998. ASTM International Directory of
          Testing Laboratories.

5      American Council of Independent Laboratories, Inc. 1992-1993. Directory: A Guide to
          Leading Independent Testing, Research, and Inspection Firms in America.

6      DynCorp. 1996. Directory of Environmental Testing Laboratories.

1      Code of Federal Regulations Title 40 Part 136—Guidelines Establishing Test Procedures
          for the Analysis of Pollutants

8      Code of Federal Regulations Title 40 Part 141—National Primary Drinking Water
          Regulations, Subpart C—Monitoring and Analytical Requirements

9      U.S. EPA. September 1994. NPDES Compliance Inspection Manual, EPA 300-B-94-
       014,
          Chapter 4 (laboratory procedures).
                                         41

-------
              APPENDIX A




ANALYTICAL SERVICES REQUEST FORM EXAMPLE

-------
                           Analytical Services Request
Client Name:
Point of Contact (name, telephone and fax number, and email address):


Date of Request:


1.      General description of analytical services requested:

2.      Definition and number of samples involved (specify wastewater, groundwater, sludge,
       soil, etc.):

3.      Purpose of analysis (NPDES, SDWA, RCPvA compliance monitoring, etc.):

4.      Estimated date(s) of sample collection:

5.      Estimated date(s) and method of shipment:

6.      Sampling/shipping contact (name and telephone number):

7.      Holding times associated with analysis (specify number of days, or state "per method"):

8.      Number of days after sample receipt that data are required:

9.      Analytical method required (specify method number, source, and date, and attach copy
       where practical):

10.    Special technical instructions (provide information on known problems, possible solutions,
       matrix effects, etc.):

11.    Data reporting requirements (specify format of data, QA/QC reports, number of copies,
       etc.):

12.    Sensitivity required (specify "per requested method," or list analyte names, CAS numbers,
       and quantitation limits required):

13.    Quality control requirements (summarize QC operations specified in the referenced
       method, and any additional requirements):

14.    Action required if QC limits exceeded (specify reanalysis, contacting client immediately,
       etc.):

15.    Other (use additional sheets or attach supplementary information, as needed):

-------
   APPENDIX B




BID SHEET EXAMPLE

-------
    Bid Response for Analysis of Effluent and Marine Water Samples for Trace Metals
Laboratory name:
Laboratory contact:
Phone Number:
Bid Deadline (Day, Date, Time)
Friday, August 29, 1997, 5:00 pm EST
Estimated Award Date:
September 17, 1997
Data deliverables due within 35 calendar days from
receipt of last sample at lab
Liquidated damages will be assessed at a rate of 2% of the
per-sample cost for each day that data is late
Period of performance: From the date your bid price is
accepted until June 13, 1998
Bid prices listed below shall be valid for a period of 90
calendar days from the bid deadline date.
Parameter
Method

1631
1631

1632
1632

1636

1637
1637

1638

1639

1640
1640
Matrix

POTW effluent
Marine Water

POTW effluent
Marine Water

POTW effluent

POTW effluent
Marine Water

POTW effluent

POTW effluent

POTW effluent
Marine Water
Analyses
Field
Samples

14
4

14
4

12

14
4

14

28

14
4
MS

1
1

1
1

1

1
1

1

1

1
1
(A) (B) (A x B)
Total
Analyses

15
5

15
5

13

15
5

15

29

15
5
Cost per
Analysis


















Total
Project Cost
Total
Cost



















Note to bidding laboratories:
Bid sheets must be accompanied by the following prequalification criteria:
1. Method blank analysis results for each method.
2. Method detection limit study for each method and each matrix set (POTW and marine water).
  water) on which you submit a bid.
EPA will evaluate these  data to identify laboratories qualified to participate in the study.	

-------
               APPENDIX C




GENERAL LABORATORY AUDIT CHECKLIST EXAMPLE

-------
                        General Laboratory Audit Checklist
Laboratory:
Audit dates:
Audit team:
Section 1:    Quality Assurance Management Systems

1.      Is there a quality assurance program plan (or equivalent) for the contract under which
       this work is being performed?
2.    Are the staff familiar with the plan?
Strengths and weaknesses:
Draft                                    C-l

-------
Section 2:   Project Management Systems

1.     Does a QAPP (or equivalent) exist for the project or work assignment?

2.     Are the staff familiar with the plan?

3.     Are all the specific elements in the QAPP included in the laboratory's QA plan?

4.     Has the QAPP (or equivalent) been approved?

5.     Have the  requirements set forth in the QA plan been met?

6.     Are deliverables on  time?

7.     Is sufficient coordination occurring between the laboratory and client project managers?

8.     Are project files available?

9.     Are software packages used by the laboratory for data reduction adequately described
       in project  files?


Strengths and weaknesses:
Draft                                     C-2

-------
Section 3:   Laboratory Management Systems

1.     Has sufficient laboratory space been allocated?

2.     Have contamination-free areas been provided for trace-level work?

3.     Are reagent-grade or higher purity chemicals used to prepare standards?

4.     Is the following information documented for all reagents/standards used?

       a.     Manufacturer

       b.     Date of receipt

       c.     Date opened

       d.     Purity

       e.     Lot number

5.     Are notebooks being kept in accordance with good laboratory practice? Are laboratory
       notebooks controlled?

6.     Have standard operating procedures (SOPs) been written where appropriate?

7.     Do staff have copies of current SOPs? Are SOPs controlled documents?

8.     Are staff performing operations according to SOPs?

9.     Does documentation exist for standards preparation  that uniquely identifies the
       reagents/solvents used and the method of preparation?

10.    Does documentation exist for identification of standard preparer and date of standard
       preparation?

11.    Are calibration standards validated prior to use?

12.    Are standards replaced at the proper intervals?

13.    Are samples subject to a chain-of-custody system?

       a.     Are they uniquely identified?

       b.     Is their storage documented and inventoried?

14.    Are manufacturers' maintenance manuals available?

15.    Are maintenance logs  kept for lab equipment/instrumentation?

16.    Is service on equipment instrumentation readily available?


Draft                                     C-3

-------
 17.    Are replacement parts for equipment/instrumentation available?

 18.    Is the analytical balance located in an area free of drafts and rapid temperature
       changes?

 19.    Do balances have calibration stickers showing date of last certified calibration and date
       of next scheduled calibration?

 20.    Are records available for in-house calibration/checking balances?

 21.    Do micropipettes have logs indicating calibration checks performed in-house?

 22.    Do records exist for monitoring of laboratory water systems?

 23.    Is everyone aware of disposal plan? Is it adhered to?

 24.    Are glassware cleaning procedures adequate?

 25.    Are temperature logs available for freezers?

 26.    Are certified material standards used for all parameters  such as material is available
       for?

 Strengths and weaknesses:
Draft                                      C-4

-------
Section 4:   Data Management Systems

1.     Are entries to logbooks signed, dated and legible?

2.     Are changes to logbooks dated and initialed by the person who made them?

3.     Can data be tracked from the project files?

4.     Do the project files identify the specific pieces of instrumentation that were used?

5.     Have lab data management systems been validated prior to use?

6.     Are data manipulation procedures adequately described?

7.     Are data (electronic and hardcopy) archived in a retrievable fashion?

8.     Is there a projection/run tracking/filing system in place?

9.     Is it possible to back-track and validate a final piece of data from it's beginning?

10.    Are data periodically confirmed by independent (i.e. manual) reduction?

11.    Are there written instructions for data receipt, storage, retrieval?

12.    Are documents issued by the work assignment subject to a document control system?

13.    Are data entered into the computer "checked at  least three times by at least two
       people?

14.    Are lab notebooks inspected by the group leader?

15.    Is the inspection documented?

Strengths and weaknesses:
Draft                                     C-5

-------
Section 5:   Problem Resolution



1.     Has a person been designated to follow-up on previously identified problems?



2.     Has a time frame been stipulated for resolving problems?



3.     Does documentation of the resolution of problems exist?




Strengths and weaknesses:
Draft                                    C-6

-------
         APPENDIX D




DATA REVIEW CHECKLIST EXAMPLE

-------
                                     Data Review Checklist
          The following pages contain a data review checklist that may be used by data reviewers,
         laboratory personnel, and other parties to document the results of each data inspection in
                                       a standardized format.
Draft                                          D-l

-------
                                               Data Inspection Checklist
                                                  Summary Information
   1.  Name of Reviewer:
                                                              Title:
                       Required Samples
                                      Sample Results Provided
        Sample Location or
            Sample ID
Analyte(s)
Sample Location or
    Sample ID
Analyte(s)
   2.  Method Used:

   3.  Total No. of analytical shifts per instrument (determined from analysis run log):
            Instrument
   4.  Total No. of CCVs Required:
   (one for each 10 samples after the
   first 10 samples on each instrument)

   5.  Total No. of CCBs Required:
   (one for each CCV)

   6.  Total No. of Field Blanks Required:
   (one per site or per 10 samples, whichever
   is more frequent)

   7.  Total no. of Lab Blanks Required:
   (one per batch' per method/instrument)

   8.  Total no. of OPR analyses Required:
   (one per batch per method/instrument)

   9.  Total no. of MS/MSD samples Required:
   (one perl 0% per matrix per site)

   10. Total no. Field Duplicates Required:
   (one per 10 samples per site)

   11. Total no. of MDL results required:
   (one per method and per analvtc)	
                                                             No. of Shifts
                    Total No. of CCVs Reported:



                    Total No. of CCBs Reported:


                    Total No. of Field Blanks Reported:



                    Total No. of Lab Blanks Reported:


                    Total No. of OPR Analyses Reported:


                    Total No. of MS/MSD samples Reported:


                    Total No. of Field Duplicates Reported:


                    Total No. of MDL Results Reported:
Draft
                D-2

-------
   15.                                      Initial Precision and Recovery (IPR)




   a.        Were IPR data reported for each analyte?                                             Dyes     Duo




   b.        Did all IPR aliquots meet required recovery criteria (x)?                                 Dyes     Dno




   c.        Did the standard deviation (s) of each IPR scries meet the required criterion?              Dyes     Dno




   d.        If any item in a - c above was answered "no", document problem below.




            Analvtc    Ave. Result Reported (X)           RSD Reported     Affected Samples
   16.                                     Ongoing Precision and Recovery (OPR)




   a.        Were OPR data reported for each analyte, instrument, and batch?                        Dyes     Dno




   b.        Did all OPR samples meet required recovery criteria (x)?                                Dyes     Dno




   c.        If item a or b above was answered "no", document problem below.




            Annlvtc     OPR Recovery fX) Reported         Shifts Missing OPR          Affected Samples
   17.                    Continuing Calibration Verification (CCV)/Continuing Calibration Blank (CCB)




   a.        Were CCVs run prior to each batch of 10 samples on each instrument?                   Dyes     Dno




   b.        Were all CCV results within the specified windows?                          Dyes     Dno




   c.        Was each CCV followed by a CCB?                                                  Dyes     Dno




   d.        Was each CCB free from contamination?                                   Dyes     Dno




   c.        If any item in a - d above was answered "no", list problems below:




            Analvtc     Affected Samples       Shift Missing CCV/CCB      Failed CCV/CCB ID
Draft                                                    D-4

-------
   18.                                         Laboratory (Method) Blanks

   a.        Was a method blank analyzed for each instrument & sample batch?                      Dyes     Dno

   b.        Was each method blank demonstrated to be free from contamination?                             Dyes     Dno

   c.        If the answer to item a or b was "no", document problems below.

            Annlvte     Affected Samples       Blank Concentration Reported    Shift Missing MB
   19.                                                 Field Blanks

   a.        Was a field blank analyzed for each 10 samples per site?                                dyes     duo

   b.        Was each field blank demonstrated to be free from contamination?                       Dyes     Dno

   c.        If the answer to item a or b was "no", document problems below.

            Analvtc     Affected Samples      Blank Concentration Reported    Shift Missing I'B
   20.                                               MS/MSD Results


   a.        Were appropriate number of MS/MSD pairs analyzed?                                 Dyes     Dno

   b.        Were all MS/MSD recoveries within specified windows?                               Dyes     Dno

   c.        Were all RPDs within the specified window?                                                   Dyes    Dno

   d.        Was appropriate corrective action (e.g., MSA for GFAA, serial dilution
            for ICP) employed on affected samples?                                              Dyes     Dno

   c.        If the answer was "no" to items a - d above, document affected samples:


            Analvte     MS % R        MSP % R         MS/MSD RPD     Affected Samples
   21.                                            Additional Information


   a.       Were Instrument Tune Data Provided?                                              Dyes    Dno

   b.       Were equipment blanks demonstrated to be free from contamination?                            Dyes    Dno

   c.       Were statements of data quality provided?                                           Dyes    Dno

   d.	Did field duplicate demonstrate acceptable precision?	Dyes    Dno
Draft                                                   D-5

-------
 REGION VI LIBRARY
 U. S. ENVIRONMENTAL PROTECTION
  AGENCY
 1445 ROSS AVENUE  ??
 DALLAS, TEXAS  7520?
EPA/CN-68-C3-0337
Guide to laboratory contracting-
July 1998

-------