Section No. 3.9
Revision No. 0
Date January 4, 1962
Page 1 .of 10
Section 3.9
METHOD 13B - DETERMINATION OF TOTAL FLUORIDE
EMISSIONS FROM STATIONARY SOURCES
(Specific-Ion Electrode Method)
OUTLINE
Number of.
Section Documentati on Pages
SUMMARY 3.9 1
METHOD HIGHLIGHTS .3.9 8
METHOD DESCRIPTION
1. PROCUREMENT OF APPARATUS
AND SUPPLIES 3.9.1 20
2. CALIBRATION OF APPARATUS 3.9.2 25
3. PRESAMPLING OPERATIONS 3.9.3 6
4. ON-SITE MEASUREMENTS 3.9.4 21
5. POSTSAMPLING OPERATIONS 3.9.5 19
6. CALCULATIONS 3.9.6 7
7. MAINTENANCE 3.9.7 3
8. AUDITING PROCEDURES 3.9.8 7
9. RECOMMENDED STANDARDS FOR
ESTABLISHING TRACEABILITY 3.9.9 1
10. REFERENCE METHOD 3.9.10 2
11. REFERENCES 3.9.11 1
12. DATA FORMS 3.9.12 22
-------
Section No. 3.9
Revision No. 0
Date January 4, 1982
Page 2 of 10
SUMMARY
In Method 13B, total fluorides (gaseous and particulate) are
extracted isokinetically from the source by using a sampling
train similar to the one specified in Method 5 (Section 3.4 of
this Handbook); however, the filter does not have to be heated,
and it may be located either immediately after the probe or
between the third and fourth impingers.
, The specific-ion electrode method for quantitatively mea-
suring the fluorides collected in the train is applicable to
fluoride (F) emissions from stationary sources, but not to fluo-
rocarbons such as Freon. The concentration range of the method
is from 0.02 to 2,000 pg F/ml; <0.1 pg F/ml requires extra care.
Sensitivity of the method has not been determined.
An interferent in the collection of fluorides is grease on
sample-exposed surfaces. The fluoride absorption into the grease
causes low results due to a lack of sample recovery. If it can
be shown to the satisfaction of the administrator that samples
contain only water soluble fluorides, fusion and distillation may
be omitted from the analysis.
-------
Section No. 3.9
Revision No. 0
Date January 4, 1982
Page 3 of 10
METHOD HIGHLIGHTS
Section 3.9 (Method 13B) describes specifications for the
sampling and analysis of total fluoride emissions from stationary
sources. A gas sample is isokinetically extracted from the
source stream, and the fluorides in the stream are collected in
the sampling train.
The sampling train is similar to that in EPA Method 5, with
a few exceptions—the filter does not have to be heated, and it
may be located either immediately after the probe or between the
third and fourth impingers. If it is between the probe and the
first impinger, a borosilicate glass or stainless steel filter
holder with a 20-mesh stainless steel screen filter support and a
silicone rubber gasket must be used. If it is between the third
and fourth impingers, a glass frit filter support may be used.
Sampling is generally the same as in Method 5, but a nozzle
size that will maintain an isokinetic sampling rate of <28 Ji/min
(<1.0 ft3/min) must be used. Samples and standards must be the
same temperature during analysis by the specific-ion electrode
(SIE). A change of 1°C (2°F) will cause a 1.5% relative error in
the sample measurements. Lack of stability in the electrometer
can also cause significant error in the results, but the main
cause of error has been found to be distillation during'sample
analyses.
The collected sample is recovered by transferring the mea-
*.
sured condensate and impinger water to a sample container, adding
the filter and the rinsings of all sample-exposed surfaces to
this container, and fusing and distilling the sample. The dis-
tilled sample is then analyzed with a SIE. Fusion and distilla-
tion may be omitted if it can be shown to the satisfaction of the
administrator that the samples contain only water soluble fluo-
rides .
Collaborative tests have shown that fluoride concentrations
from 0.1 to 1.4 |jg F/m3 could be determined with an intralabora-
tory precision of 0.037 jjg F/m3 and an interlaboratory precision
-------
Section No. 3.9
Revision No. 0
Date January 4, 1982
Page 4 of 10
of 0.056 pg F/m3. For these tests six contractors simultaneously
took duplicate samples from a stack. The collaborative test did
not find any bias in the analytical method.1
The Method Description (Sections 3.9.1 to 3.9.9) is based on
the detailed specifications in the Reference Method (Section
3.9.10) promulgated by EPA on June 20, 1980.2
1. Procurement of Apparatus and Supplies
Section 3.9.1 gives specifications, criteria, and design
features for the required equipment and materials. The sampling
apparatus for Method 13B has the same design features as that of
Method 5, except for the positioning of the filter in the sam-
pling train. This section can be used as a guide for procurement
and initial checks of equipment and supplies. The activity
matrix (Table 1.1) at the end of the section is a summary of the
details given in the text and can be used as a quick reference.
2. Pretest Preparations
Section 3.9.2 describes the required calibration procedures
for the Method 13B sampling equipment (same, as Method 5), except
for the SIE. A pretest checklist (Figure 3.1 or a similar form)
should be used to summarize the calibration and other pertinent
pretest data.
Section 3.9.3 describes the preparation of supplies and
equipment needed for the sampling. The pretest preparation form
(Figure 3.2 of Section 3.4.3) can be. used as an equipment check-
list. Suggestions for packing the equipment and supplies for
shipping are given to help minimize breakage.
Activity matrices for the calibration of equipment and the
presampling operations (Tables 2.1 and 3.1) summarize the activi-
ties detailed in the text.
3. On-site Measurements
Section 3.9.4 describes procedures for sampling and sample
recovery. A checklist (Figure 4.5) is an easy reference for
field personnel to use in all sampling activities.
-------
Section No. 3.9
Revision No. 0
Date January 4, 1982
Page 5 of 10
4. Posttest Operations
Section 3.9.5 describes the postsampling activities for
checking the equipment and the analytical procedures. A form is
given for recording data from the posttest equipment calibration
checks; a copy of the form should be included in the emission
test final report. A control sample of known (F) concentration
should be analyzed before analyzing the sample for a quality
control check on the analytical procedures. The detailed ana-
lytical procedures can be removed for use as easy references in
the laboratory. An activity matrix (Table 5.1) summarizes the
postsampling operations.
Section 3.9.6 describes calculations, nomenclature, and
significant digits for the data reduction. A programmed calcu-
lator is recommended to reduce calculation errors.
Section 3.9.7 recommends routine and preventive maintenance
programs. The programs are not required, but their use should
reduce equipment downtime.
5. Auditing Procedures
Section 3.9.8 describes performance and system audits.
Performance audits for both the analytical phase and the data
processing are described. A checklist (Figure 8.2) outlines a
system audit.
Section 3.9.9 lists the primary standards to which the
working standards or calibration standards should be traceable.
6. References
Section 3.9.10 contains the promulgated Reference Method;
Section 3.9.11 contains the references used throughout this
text; and Section 3.9.12 contains copies of data forms recom-
mended for Method 13B.
-------
Section No. 3.9
Revision No. 0
Date January 4, 1982
Page 6 of 10
PRETEST SAMPLING CHECKS
(Method 13B, Figure 3.1)
Date Calibrated by
Meter box number AH@
Dry Gas Meter*
Pretest calibration factor Y | (within ±2% of the
average factor for each calibration run).
Impinger Thermometer
Was a pretest temperature correction used? _. _._ yes no
If yes, temperature correction (within ±1°C (2°F)
of reference values for calibration and within ±2°C (4°F) of
reference values for calibration check)
Dry Gas Meter Thermometers
Was a pretest temperature correction made? __] yes .no
If yes, temperature correction (within ±3°C (5.4°F) of
reference value for calibration and within 6°C (10.8°F) of
reference values for calibration check)
Stack Temperature Sensor*
Was a stack temperature sensor calibrated against a reference
thermometer? yes ,___ no
If yes, give temperature range with which the readings agreed
within ±1.5% of the reference values to K (°R)
Barometer
Was the pretest field barometer reading correct? yes no
(within ±2.5 nun (0.1 in.) Hg of the mercury-in-glass barometer)
Nozzle*
Was the nozzle calibrated to the nearest 0.025 mm (0.001 in.)?
yes no
*Most significant items/parameters to be checked.
-------
Section No. 3.9
Revision No. 0
Date January 4, 1982
Page 7 of 10
ON-SITE MEASUREMENTS
(Method 13B, Figure 4.5)
Apparatus
Probe nozzle: stainless steel glass
Button-hook elbow size
Clean?
Probe liner: borosilicate quartz other
Clean?
Heating system*
Checked?
Pitot tube: Type S other
Properly attached to probe?*
Modifications
_Pitot tube coefficient
Differential pressure gauge:two inclined manometers
other ^ sensitivity
Filter holder:borosilicate glass glass frit
filter support silicone gasket other
Clean? "
Condenser: number of impingers
Clean?
Contents: 1st 2nd 3rd 4th
Cooling system
Proper connections?
Modifications
Barometer: mercury aneroid other
Gas density determination: temperature sensor type
pressure gauge
temperature sensor properly attached to probe?*
Procedure
Recent calibration: pitot tubes*
meter box* thermometers/thermocouples1
Filters checked visually for irregularities?*
Filters properly labeled?*
Sampling site properly selected?
Nozzle size properly selected?*
Selection of sampling time?
All openings to sampling train plugged to prevent pretest con
tamination?
Impingers properly assembled? -••
Filter properly centered?
Pitot tube-lines checked for plugging or leaks?*
(continued)
-------
Section No. 3.9
Revision No. 0 /"""N
Date January 4, 1982
Page 8 of 10
Figure 4.5 (continued)
Meter box leveled? Periodically?
Manometers zeroed?
AH@ from most recent calibration
Nomograph setup properly?
Care taken to avoid scraping nipple or stack wall?*
Effective seal around probe when in-stack?
Probe moved at proper time?
Nozzle and pitot tube parallel to stack wall at all times?*
Filter changed during run?
Any particulate lost?
Data forms complete and data properly recorded?*
Nomograph setting changed when stack temp changed significantly?
Velocity pressure and orifice pressure readings recorded
accurately?*
Sampling performed at a rate less than 1.0 cfm
Posttest leak check performed?* (mandatory)
Leakage rate @ in. Hg
Orsat analysis from stack integrated ' /"~*\
Fyrite combustion analysis sample location V_x
Bag system leakchecked?* _i
If data forms cannot be copied, record:
approximate stack temp volume metered
% isokinetic calculated at end of each run __
SAMPLE RECOVERY
Brushes: nylon bristle other
Clean?
Wash bottles: polyethylene or glass
Clean?
Storage containers: polyethylene other
Clean? Leakfree?
Graduated cylinder/or balance: subdivisions <2 ml?*
other
Balance: type
Probe allowed to cool sufficiently?
Cap placed over nozzle tip to prevent loss of particulate?*
During sampling train disassembly, are all openings capped?
Clean-up area description:
Clean? Protected from wind?
Filters: paper type
_
Silica gel: type (6 to 16 mesh)? new? _ used?
Color? Condition?
Filter handling: tweezers used?
surgical gloves? other
Any fluoride spilled?*
(continued)
-------
Section No. 3.9
Revision No. 0
Date January 4, 1982
Page 9 of 10
Figure 4.5 (continued)
Water distilled?
Stopcock grease: acetone-insoluble?
heat-stable silicone? ~
Probe handling: distilled water rinse
Fluoride recovery from: probe nozzle
other
probe fitting
front half of filter holder
Blank: filter
probe liner
distilled water
Any visible particles on filter holder inside probe?:*
All jars adequately labeled? __
Liquid level marked on jars?*
Locked up?
Filter blank
Sealed tightly?
*M6st significant items/parameters to be checked.
-------
Section No. 3.9
Revision No. 0
Date January 4, 1982
Page 10 of 10
o
METHOD 13B CHECKLIST FOR AUDITORS
(Method 13B, Figure 8.2)
Yes
No
Comment
OPERATION
Presampling Preparation
1. Knowledge of process conditions
2. Calibration of equipment, before each
field test
On-Site Measurements
3. Sample train assembly
4. Pretest leak check
5. Isokinetic sampling
6. Posttest leak check
7. Record process conditions during sample
collection
8. Sample recovery and data integrity
Postsampling
9. Accuracy and precision of control sample
analysis
10. Recovery of samples for distillation
11. Calibration checks
12. Calculation procedure/check
O
General Comments:
O
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 1 of 20
METHOD DESCRIPTION
1.0 PROCUREMENT OF APPARATUS AND SUPPLIES
A schematic of the sampling train used in Method 13B is
shown in Figure 1.1. The train and the sampling procedures are
similar to EPA Method 5; the procedures and equipment for Methods
13A and 13B are identical. Commercial models of the train are
available. For those who want to build their own, construction
details are in APTD-0581;3 allowable modifications are described
herein. The operating, maintenance, and calibration procedures
for the sampling train are in APTD-0576.4 Since correct usage is
important in obtaining valid results, all users are advised to
read this document and to adopt its procedures unless alterna-
tives are outlined herein.
Specifications, criteria, and/or design features are given
in this section to aid in the selection of equipment which as-
sures collection of good quality data. Procedures and limits
(where applicable) for acceptance checks are also given.
During procurement of equipment and supplies, a log (Figure
1.2) should be used to record the descriptive titles and identi-
fication numbers (if applicable) of the equipment and the results
of the acceptance checks; a blank copy of the procurement log is
in Section 3.9.12 for the convenience of the Handbook user. If
calibration is required for the acceptance check, a calibration
log should be used to record the data. Table 1.1 at the end of
this section summarizes the quality assurance activities for the
procurement and acceptance of apparatus and supplies.
1.1 Sampling Apparatus
1.1.1 Probe Liner - The sampling probe should be constructed of
borosilicate glass (Pyrex) or 316 stainless steel tubing with an
outside diameter (OD) of about 16 mm (0.625 in.); it should be
encased in a stainless steel' sheath with an OD of 25.4 mm (1.0
in. ).
(I It-'
-------
1.9-Z.5 cm
(0.75-1 in.)
1.9 cm(0.75 1n.)
PITOT TUBE
TEMPERATURE
SENSOR
7
PROBE
, OPTIONAL
'FILTER HOLDER'
I LOCATION I
!^U!LJ FILTER HOLDER
PROBE
TYPE S C
PITOT TUBE
^THERMOMETER
CHECK
-.VALVE
ORIFICE
MANOMETER
DRY TEST
METER
Figure 1.1. Fluoride sampling train.
VACUUM
LINE
*u \j %j (/)
a> ru n> CD
*> rt < o
fD (D (-•• f+
tn H-
H-O
to
VD
00
fO
O
o
o
-------
Item description
Quantity
Purchase
order
number
Vendor
Date
Ordered
Received
Cost
Dispo-
sition
Comments
C0
n>
ja en
n> o>
rt < o
0) H-
M-O
§d
CO
Figure 1.2. Example of a procurement log.
vo
oo
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 4 of 20
A heating system may be required to maintain the exit gas at
120° ±14°C (248° ±25°F) during sampling. Other temperatures may
be specified by a subpart of the regulations or approved by the
administrator. Since the probe outlet temperature is not usually
monitored during sampling, probes constructed in accordance to
APTD-0581s and calibrated with procedures in APTD-05764 will be
acceptable.
Upon receiving a new probe, visually check for specifica-
tions (i.e., the length and composition ordered) and for breaks
or cracks; leak check on a sampling train (Figure 1.1); and check
the nozzle-to-probe connection with a Viton-O-ring or with Teflon
ferrules for glass liners or stainless steel ferrules for stain-
less steel liners.
The probe heating system should be checked as follows:
1. Connect the probe by attaching the nozzle to the pump
inlet.
2. Connect the probe heater to the electrical source, and
turn it on for 2 or 3 min; it should become warm to the touch.
3. Start the pump, and adjust the needle valve until a
flow rate of about 0.02 m3/min (0.75 ft3/min) is achieved.
4. Be sure that the probe remains warm to the touch and
that the heater maintains the exit gas at a minimum of 100°C
(212°F); if not, repair, return to the supplier, or reject the
probe.
1.1.2 Probe Nozzle - The probe nozzle should be designed with a
sharp, tapered leading edge and should be constructed of either
seamless 316 stainless steel tubing or glass formed in a button-
hook or elbow configuration. The tapered angle should be £30°,
with the taper on the outside to preserve a constant inside
diameter (ID).
A range of nozzle ID's [e.g., 0.32 to 1.27 cm (0.125 to 0.5 {)
in.) in increments of 1.6 mm (0.0625 in.)] should be available
for isokinetic sampling. Larger nozzle sizes may be required
-------
Section No. 3.'9.1
Revision No. 0
Date January 4, 1982
Page 5 of 20
for hi-vol sampling trains or for very low stack gas velocities.
Each nozzle should be engraved with an identification number for
inventory and for calibration purposes.
Upon receipt of the nozzle from the manufacturer and before
each test, inspect it for roundness and corrosion and for damage
(nicks, dents, and burrs) to the tapered edge, and check the ID
with a micrometer. (Calibration procedures are in Section
3.9.2.) Slight variations from exact ID'S should be expected due
to machining tolerances. Reshape, return to supplier or reject.
1.1.3 Pitot Tube - The pitot tube, preferably of Type S design,
shown in Figure 1.1 should meet the requirements of Method 2
(Section 3.1.2). Proper pitot-tube-sampling-nozzle configuration
for prevention of aerodynamic interference is shown in Figures
2.6 and 2.7 of Method 2 (Section 3.1.2).
Visually inspect the vertical and horizontal tip alignments.
If the tube is purchased as an integral part of a probe assembly,
check the dimensional clearances using Figures 2.6 and 2.7 of
Section 3.1.2. Repair or return any pitot tube that does not
meet specifications.
1.1.4 Differential Pressure Gauge - The differential pressure
gauge should be an inclined manometer or the equivalent, as
specified in Method 2, Section 3.1.2. Two gauges are required.
One is used to monitor the stack velocity pressure, and the other
to measure the orifice pressure differential.
Initially, check the gauges against a gauge-oil manometer at
a minimum of three points—0.64, 12.7, and 25.4 mm (0.025, 0.5,
and 1.0 in.) H20—to see if they read within 5% at each test
point. Repair or return to the supplier any gauge that does not
meet these requirements.
1.1.5 Filters - If the filter is between the third and fourth
impingers, use a Whatman No. 1 (or equivalent) filter, sized to
fit the filter holder. If it is between the probe and the first
impinger, use any suitable medium (e.g., paper or organic mem-
brane ) as long as the filter can withstand prolonged exposure up
-------
Section No. 3.9.1
Revision No. 0 S~~\
Date January 4, 1982 ( )
Page 6 of 20 V^X
to 135°C (275°F) and has >_95% collection efficiency (£5% penetra-
tion) for 0.3-p dioctyl phthalate smoke particles.
Conduct the filter efficiency test before beginning the
sampling by using either the ASTM Standard Method D2986-71 or
test data from the supplier's quality control program. The
filter should have a low F blank value (<_0.015 mg F/cm2 of filter
area);' determine the average values of at least three filters
from the lot to be used for sampling. Glass fiber filters gen-
erally have high and/or variable F blank values, and thus are not
acceptable.
1.1.6 Filter Holder - If the filter is located between the probe
and first impinger a borosilicate glass or stainless steel filter
holder with a 20 mesh stainless steel mesh frit filter support
and a silicone rubber gasket is required by the Reference Method. -^
If it is between the third and fourth impingers, the tester may ( )
use borosilicate glass with a glass frit filter support and a
silicone rubber gasket. Other gasket materials (e.g., Teflon or
Viton) may be used if approved by the administrator.
The holder design must provide a positive seal against
leakage from the outside or around the filter. The holder should
be durable, easy to load, and leak free in normal applications.
Check visually before use. If immediately following the probe,
the filter should be positioned toward the flow.
1.1.7 Filter Heating System - Any heating system may be used
which is capable of maintaining the filter holder at 120° ±14°C
(248° ±25°F) during sampling. Other temperatures may be speci-
fied by a subpart of the regulations or approved by the admini-
strator. The heating element should be easily replaceable in
case of malfunction during sampling. A gauge capable of mea-
suring within 3°C (5.4°F) should be used to monitor the tempera-
ture around the filter during sampling.
Check the heating system and the temperature monitoring
device before sampling.
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 7 of 20
1.1.8 Condenser - Four impingers with leak-free, noncontaminated
ground-glass or similar fittings should be connected in series.
The first, third, and fourth impingers must be £ modified
Greenburg-Smith design that has a glass tube (instead of inserts)
with an unconstricted 13-mm (0.5-in.) ID extending to within 13
mm (0.5 in.) of the flask bottom. The second impinger must be a
Greenburg-Smith with the standard tip and plate. Modifications—
for example, using flexible connections between impingers, using
materials other than glass, or using a flexible vacuum hose to
connect the filter holder to the condenser—must be approved by
the administrator." The fourth impinger outlet connection must
allow insertion of a thermometer capable of measuring ±1°C (2°F)
of true value in the range of 0° to 25°C (32° to 77°F).
Alternatively, any system that cools the gas stream and
allows measurement of the condensed water and the water vapor
leaving the condenser, each to within 1 ml or 1 g, may be used
with approval from the administrator.
Upon receipt of a standard Greenburg-Smith impinger, fill
the inner tube with water; if the water does not drain through
the orifice in 6 to 8 s or less, replace or enlarge the impinger
tip to prevent an excessive pressure drop in the sampling system.
Check each impinger visually for damage—breaks, cracks, or
manufacturing flaws such as poorly shaped connections.
1.1.9 Metering System - The metering system should consist of a
vacuum gauge, a vacuum pump, thermometers capable of measuring
±3°C (5.4°F) of true value in the range of 0° to 90CC (32° to
194°F), a dry gas meter with ±2% accuracy at the required sam-
pling rate, and related equipment shown in Figure 1.1. Other
metering systems capable of maintaining rates within 10% of
isokinetic and determining sample volumes within 2% may be used
if approved by the administrator. Sampling trains with metering
systems designed for rates higher than those described in APTD-
0581s and APTD-05764 may be used if the above specifications can
be met.
-------
Section .No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 8 of 20
When the metering system is used with a pitot tube, the
system should permit verification of an isokinetic sampling rate
with a nomograph or by calculation.
Upon receipt or after construction of the equipment, perform
both positive and negative pressure leak checks before beginning
the calibration procedure (Section 3.9.2). Adjust, repair, or
replace the malfunctioning item. Reject a dry gas meter if it
behaves erratically or if it cannot be adjusted. Reject the
thermometer if unable to calibrate.
1.1.10 Barometer - A mercury, aneroid, or other barometer capa-
ble of measuring atmospheric pressure to within ±2.5 mm (0.1 in.)
Hg is require'd.
Check a new barometer against a mercury-in-glass barometer
or the equivalent. In lieu of this, obtain the absolute barom-
etric pressure from a nearby weather service station and adjust
it for the elevation difference between the station and the sam-
pling point; accordingly, either subtract 2.5 mm Hg/30 m (0.1 in.
Hg/100 ft) from the station value for an elevation increase or
add the same for an elevation decrease. If the barometer cannot
be adjusted to agree within 2.5 mm (0.1 in.) Hg of the reference
barometric pressure, either return it to the manufacturer or
reject it.
1.1.11 Gas Density Determination Equipment - A temperature sen-
sor and a pressure gauge (Method 2, Section 3.1.2) are required.
A gas analyzer (Me'thod 3, Section 3.2.2) may be required.
The temperature sensor should be permanently attached to
either the probe or the pitot tube; in either case, a fixed
configuration (Figure 1.1) should be maintained. Alternatively,
the sensor may be attached just before field use (Section 3.9.2).
1-2 Sample Recovery Apparatus
1.2.1 Probe Liner and Nozzle Brushes - Nylon bristle brushes
with stainless steel wire handles are recommended. The probe ( J
brush must be at least as long as the probe. A separate,
smaller, and very flexible brush should be used for the nozzle.
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 196.2
., Page 9 of 20
Visually check for damage upon receipt, and replace or
return to supplier if defective.
1.2.2 Wash Bottles - Two 500-ml wash bottles are recommended for
the probe and the glassware rinsings. Glass or polyethylene are
acceptable.
1.2.3 Sample Storage Containers - Recommended are 500 ml or 1000
ml chemically resistant, high-density polyethylene bottles for
storage of samples. The bottles must have leak proof screw caps
with leak proof, rubber-backed Teflon cap liners, or they must be
constructed to preclude leakage and to resist chemical attack;
wide-mouthed bottles are easiest to use.
Prior to field use, inspect the cap seals and the bottle cap
seating surfaces for chips, cuts, cracks, and manufacturing
deformities which would allow leakage.
1.2.4 Graduated Cylinder and/or Triple Beam Balance - Either a
250-ml glass (Class A) graduated cylinder or a triple beam
balance may be .used to measure the water condensed in the imping-
ers during sampling. The graduated cylinder may also be used to
measure water initially placed in the first and second impingers.
In either case, the required accuracy is 1 ml or 1 g; therefore,
the cylinders must have subdivisions <2 ml, and the balance should
be capable of weighing to the nearest 0.1 g.
1.2.5 Plastic Storage Containers - Several airtight plastic
containers are needed for storage of silica gel.
1.2.6 Funnel and Rubber Policeman - A funnel and .a rubber po-
liceman are needed to transfer the used silica gel from the
impinger to a storage container unless silica gel is weighed in
the field after the test. A Teflon policeman is helpful for
recovery of the filter. The funnel should be glass with a 100-mm
diameter and a 100-mm stem.
Visually check on receipt, and replace or return if damaged.
1.3 Distilling Apparatus
The fluoride distillation setup is shown in Figure 1.3.
1.3.1 Flasks - A long-necked, round bottom 1-liter flask with
24/40 joint grindings is needed for boiling the sample solution.
-------
a
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 10 of 20
o
CONNECTING TUBE
12-rmID
'24/40
THERMOMETER
TIP MUST
EXTEND BELOW
THE LIQUID
WITH 10/30
24/40
1-liter
FLASK
BUNSEN.
BURNER
24/40
CONDENSER
250 ml
VOLUMETRIC
FLASK
O
Figure 1.3. Fluoride distillation apparatus.
O
-------
^ Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 11 of 20
Also, a 250-ml volumetric (Class A) flask is needed to receive
the condensate.
1.3.2 Thermometer - A thermometer for checking the temperature
of the sample in the boiling flask should read within ±1°C (±2°F)
of the true value in the range of 100° to 200°C (212° to 392°F).
Check it against a mercury-in-glass thermometer.
1.3.3 Adapter - An adapter should have joint grindings (inner
and outer parts) that are 24/40 at the bottom and 10/30 at the
top that will hold a thermometer, and it should have a 24/40
joint grinding (inner part) at the end of the top sidearm that
joins the connector tube.
1.3.4 Connector Tube - A tube with a standard or a medium wall
and with a 13-mm (0.5 in.) ID is needed for connecting the
adapter to the condenser.
1.3.5- Condenser - A coiled integral Graham condenser (joint
grinding 24/40) with a jacket length of 300 mm (12 in.) is needed
for condensation of the distillate.
1.4 Miscellaneous Glassware
1.4.1 Beaker - A 1500-ml glass beaker (Class A) with 5-ml sub-
divisions is needed to receive the filtered sample from container
No. 1 or No. 2.
1.4.2 Pipettes - Several volumetric pipettes (Class A)—includ-
ing 5, 10, 20, 25, 50 mi's—should be available. Record the
stock numbers, and visually check for cracks, breaks, or manu-
facturer's flaws. If irregularities are found, either replace or
return to the supplier.
1.4.3 .Volumetric Flasks - The following volumetric flasks are
needed for performing the analysis: a 50-ml glass volumetric
flask (Class A) is needed to dilute the sample aliquot to 50 ml
with TISAB (total ionic strength adjustment buffer) in determina-
tion of fluoride concentration; a l-£ glass volumetric flask
(Class A) to dilute the fused sample to volume with distilled
water; and several 100-ml polyethylene volumetric flasks to
prepare the fluoride standardizing solution.
I
-------
}' / Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 12 of 20
1.5 Reagents and other Supplies (Sampling)
Unless otherwise indicated, all reagents should meet the
specifications of the Committee on Analytical Reagents of the
American Chemical Society (ACS); otherwise, use the best avail-
able grade.
1.5.1 Silica Gel - Use indicating type 6-16 mesh. If previously
used, dry at 175°C (347°F) for at least 2 h before reusing. New
silica gel may be used as received.
1.5.2 Water - Deionized distilled water should conform to ASTM
specification D1193-74, Type 3. At the option of the analyst,
the KMnO^ (potassium permanganate) test for oxidizable organic
matter may be omitted if high concentrations of organic matter
are not expected.
1.5.3 Crushed Ice - Enough crushed ice is • needed around the
impingers to maintain <20°C (68°F) at the impinger silica gel
outlet in order to avoid excessive moisture loss.
1.5.4 Stopcock Grease - Acetone insoluble, heat stable silicone
grease is required unless screw-on connectors with Teflon or
similar sleeves are used.
1.6 Reagents and Supplies (Sample Recovery and Analysis)
Unless otherwise indicated, all reagents should meet the
specifications of the Committee on Analytical Reagents of the
American Chemical Society (ACS); otherwise, use the best avail-
able grade.
1.6.1 Calcium Oxide (CaO) - A reagent grade or a certified ACS
grade of CaO should contain £0.005% F.
l-6'2 Filters - Whatman No. 541 (or equivalent) filters are re-
quired for filtration of the impinger contents and preparation of
the sample for analysis.
1.6.3 Phenolphthalein Indicator - A reagent grade or a certified
ACS 0.1% phenolphthalein should be a 1:1 ethanol-water mixture.
1.6.4 Sodium Hydroxide - An ACS reagent grade (or the equiva-
lent) NaOH pellets and 5M NaOH reagent grade or ACS is needed.
1.6.5 Sulfuric Acid - An ACS reagent grade (or the equivalent)
concentrated H,SOA and 25% (v/v) reagent grade- or ACS is needed. ,-
™ •* j *~~v *""}
O/7
o
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 13 of 20
1.6.6 Total Ionic Strength Adjustment Buffer (TISAB) - To ap-
proximately 500 ml of distilled water in a l-£ beaker, add 57 ml
of glacial acetic acid, 58 g of sodium chloride, and 4 g of
cyclohexylene dinitrilotetra acetic acid (CDTA). Stir to dis-
solve. Place the beaker in a water bath until it cools. Then,
slowly add 5 M NaOH, while measuring the pH continuously with a
calibrated pH-reference electrode pair, until the pH is 5.3.
Cool to room temperature. Pour into a l-£ flask, and dilute to
volume with distilled water. Commercially prepared TISAB buffer
may be substituted for the above.
1.6.7 Fluoride Standard Solution - To prepare a 0.1M fluoride
reference solution, add 4.20 grams of reagent grade sodium fluo-
ride (NaF) to a 1-2 volumetric flask, and add enough distilled
water to dissolve it. Dilute to volume with distilled water.
The NaF must be oven dried at 110°C for at least 2 h prior to
weighing.
1.7 Analytical Equipment
1.7.1 Bunsen Burner - A Bunsen burner capable of distilling 200
ml in <15 min is required for the boiling flasks.
1.7.2 Crucible - A nickel crucible with a capacity of 75 to
100 ml is needed to evaporate the water from the sample on a hot
plate.
Upon receipt, check for cracks or manufacturing flaws as
well as for capacity. If it does not meet specifications replace
or return it to the manufacturer.
1.7.3 Hot Plate - A hot plate capable of 500°C (932°F) is re-
quired for heating the sample in a nickel crucible.
Check upon receipt and before each use for damage. Check
the heating capacity against a mercury-in-glass thermometer. If
inadequate, repair or return the hot plate to the supplier.
1.7.4 Electric Muffle Furnace - An electric muffle furnace
— " -• -" -- -"" t
capable of heating to 600°C (1112°F) is needed to fuse the sam-
ple.
Check the heating capacity against a mercury-in-glass ther-
mometer. Replace or return to the manufacturer any unit which
does not meet specifications. ('
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 14 of 20
1.7.5 Balance - A balance with a capacity of 300 g ±0.5 g is
needed to determine moisture.
Check for damage against a series of standard weights upon
receipt and before each use. Replace or return to the manufac-
turer if damaged or if it does not meet specifications.
1.7.6 Analytical Balance - An analytical balance capable of
weighing to within 0.1 mg is needed for preparation of the stan-
dard fluoride solution and the analytical reagents. Check the
balance frequently with Class S weights.
1.7.7 Constant Temperature Bath - A water bath is needed to
maintain a constant room temperature for optimum measurement *of
the sample concentration.
i.7.8 Fluoride Ion Activity-Sensing Electrodes - A fluoride ion
(F~) activity-sensing electrode is required in Method 13B for
determining of F*~ ion activity in concentrations of from 1 to
10~ mol/Ji (19,000 to 0.02 ppm). The electrode should be usable'
from a pH of 1 to 8.5 at 10~6 mol/£, up to a pH of 11 at 10"2
mol/£ F~ concentration. Due to the complexing of F~ below pH of
4 and to the limited resistance of the electrode body to certain
concentrated acids, it is usually advisable to adjust the pH of
strongly acidic samples.
Check for damage and F~ sensing accuracy with a known con-
centration upon receipt and before each use. If not suitable,
replace or return to manufacturer. Either a single junction,
sleeve type reference electrode or a combination type fluoride
ion-sensing electrode built into one unit may be used.
1.7.10 Electrometer - Either a pH meter with a millivolt scale
capable of ±0.1-mV resolution or a ion meter made especially for
specific-ion use is needed to read the ion activity or the F~
concentration.
1.7.11 Magnetic Stirrer - A magnetic stirrer and TFE* fluorocar-
bon-coated stirring bars are needed for uniform mixing of the
sample solution. /""*\
1.7'. 12 Stopwatch or Clock - A stopwatch or a clock is. needed to V_y
check"minimum immersion time of electrode in sample.
o
*Mention of any trade name or specific product does not consti-
tute endorcement by the Environmental Protection Agency.
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 15 'of 20
TABLE 1.1 ACTIVITY MATRIX FOR PROCUREMENT OF APPARATUS
AND SUPPLIES
Apparatus
Sampling
Probe liner
Probe nozzle
Pitot tube
'Differential
pressure
gauge (in-
clined ma-
nometer)
Filters
Acceptance limits
Specified material of
construction; equipped
with heating system
capable of maintaining
120°±14°C (248° ±25°F)
at the exit
Stainless steel (316)
with sharp, tapered
angle <30°; differ-
ence in measured diam-
eters £0.1 mm (0.004
in.); no nicks, dents,
or corrosion
Type S (Meth 2, Sec
3.1.2); attached to
probe with impact
(high pressure) opening
plane even with or
above nozzle entry
plane
Meets- criteria (Sec
3.1.2); agrees within
5% of gauge-oil
manometer
Capable of withstand-
ing temperatures to
135°C (275°F), 95%
collection efficiency
for 0.3 urn particles,
low F blank (<0.015
mg F/cm2) ~~
Frequency and method
of measurements
Visually check the
probe and run the
heating system
Visually check upon
receipt and before
each test; use a mi-
crometer to measure
ID before field use
after each repair
Visually check for
vertical and hori-
zontal tip alignments;
check the configura-
tion and the clear-
ances; calibrate
(Sec 3.1.2, Meth 2)
Check against a gauge-
oil manometer at a
minimum of three
points: 0.64(0.025);
12.7 (0.5); 25.4(1.0)
mm (in.) H20
Check each batch for
F blank values,
visibly inspect for
pin holes or flaws
Action if
requirements
are not met
Repair, return
to supplier,
or reject
Reshape and
sharpen, re-
turn to the
supplier, or
reject
Repair or re-
turn to sup-
plier
As above
Reject batch
(continued)
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 16 of 20
o
TABLE 1.1 (continued)
Apparatus
Filter holder
Condenser
Vacuum gauge
Vacuum pump
Barometer
Orifice meter
Dry gas meter
(continued)
Acceptance limits
Leak free; borosilicate
glass
Four impingers, standard
stock glass; pressure
drop not excessive
0-760 mm (0-30 in.) Hg,
±25 mm (1 in.) at
380 mm (15 in.) Hg
Leak free; capable of
maintaining flow rate
of 0.02-0.03 mVmin
(0.7 to 1.1 ftVmin)
for pump inlet vacuum
of 380 mm (15 in.) Hg
Capable of measuring
atmospheric pressure
±2.5 mm (0.1 in.) Hg
AH@ of 46.74± 6.35 mm
(1.84 ± 0.25 in.) H20
at 20°C (68°F);
optional
Capable of measuring
volume within ±2% at a
flow rate of 0.02
tnVmin (0.7 ftVmin)
Frequency and method
of measurements
Visually check before
use
Visually check upon
receipt; check pres-
sure drop
Check against mer-
cury U-tube manometer
upon receipt
Check upon receipt
for leaks and capaci-
ty
Check against a mer-
cury-in-glass barom-
eter or equivalent;
calibrate (Sec 3.1.2)
Upon receipt, visual-
ly check for damage;
calibrate against wet
test meter
Check for damage upon
receipt and calibrate
(Sec 3.9.2) against
wet test meter
Action if
requirements
are not met
Return to
supplier
As above
Adjust or re-
turn to sup-
plier
Repair or re-
turn to sup-
plier
Determine cor-
rection fac-
tor, or reject
Repair or re-
turn to sup-
plier
Reject if dam-
aged, behaves
erratically,
or cannot be
properly ad-
justed
O
o
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 17 of 20-
Table 1.1 (continued)
Apparatus
Thermometers
Sample Recovery
Probe liner and
probe nozzle
brushes
Wash bottles
Storage con-
tainer
Graduated
cylinder
Funnel
Rubber police-
man
Acceptance limits
±1°C (2°F) of true
value in the range of
0° to 25°C (32° to 77°F)
for impinger thermometer
and ±3°C (S.A'F) of true
value in the range of
0°C to 90°C (32° to
194°F) for dry gas
meter thermometers
Nylon bristles with
stainless steel han-
dles; properly sized
and shaped
Polyethylene or glass,
500 ml
.High-density polyeth-
ylene, 1000 ml
Glass, Class A, 250 ml
Glass, Class A, diameter
100 mm; stem length
100 mm
Properly sized
Frequency and method
of measurements
Check upon receipt
for dents or bent
stem, and calibrate
(Sec 3.9.2) against
mercury-in-glass
thermometer
Visually check for
damage upon receipt
Visually check for
damage upon receipt
Visually check for
damage upon receipt;
be sure caps make
proper seals
Upon receipt, check
for stock number,
cracks, breaks, and
manufacturer flaws
Visually check for
damage upon receipt
Visually check for
damage upon receipt
Action if
requirements
are not met
Reject if un-
able to cali-
brate
Replace or re-
turn to sup-
plier
As above
As above
As above
As above
As above
(continued)
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 18 of 20
o
TABLE 1.1 (continued)
Apparatus
Pipettes, volu-
metric flask
beaker, flask
adapter, con-
denser, con-
nection tube,
Erlenmeyer
flask
Distallation
Apparatus
Bunsen burner
Crucible
Analytical
Equipment
Hot plate
Electric muffle
furnace
Acceptance 1imits
Glass, Class A
Capable of distilling
220 ml in <15 min
Nickel material; 75-
100 ml
Heating capacity of
500°C (932°F)
Heating capacity of
600°C
Frequency and method
of measurements
Upon receipt, check
for stock number,
cracks, breaks and
manufacturer flaws
Visually check upon
receipt; check heat-
ing capacity, check
for damage
Check upon receipt
for cracks or flaws
Check upon receipt
and before each use
for damage; check
heating capacity
against mercury-in-
glass thermometer
Check upon receipt
and before each use
for damage; check
heating capacity
upon receipt against
mercury-in-glass
thermometer
Action if
requirements
are not met
Replace or re-
turn to sup-
plier
Replace
O
Replace or re-
turn to manu-
facturer
Replace or re-
turn to manu-
facturer
Replace or re-
turn to manu-
facturer
O
(continued)
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 19 of 20
Table 1.1 (continued)
Apparatus
Balance
Water bath
Fluoride ion
activity-sen-
sing electrode
Reference
electrode
Electrometer
Reagents
Filters
Silica gel
Acceptance limits
Capacity of 300 g ±0.5g
Capable of maintaining
constant room tempera-
ture
Capable of measuring
F concentration from
1 to 10-6 mol/£
(19,000 to 0.02 ppm)
Should provide stable
output
Capable of reading to
±0.1 mV resolution with
temperature compensa-
tion
Whatman No.
equivalent
541 or
Indicating Type 6-16
mesh
Frequency and method
of measurements
Check for damage and
against series of
standard weights upon
receipt and before
each use
Check with mercury-
in-glass thermometer
Check for damage and
F sensing accuracy
with a known con-
centration upon re-
ceipt and before
each use
Check visually for
cracks or breaks
Upon receipt and
before each use,
check for per-
formance accuracy
with a_known stan-
dard F solution
Visually check for
damage upon receipt'
Upon receipt check
label for grade or
certification
Action if
requirements
are not met
Replace or re-
turn to manu-
facturer
Repair
Replace or re-
turn to manu-
facturer
Replace or re-
turn to manu-
facturer
Replace or re-
turn to manu-
facturer
Replace or re-
turn to sup-
plier
Replace or re-
turn to manu-
facturer
(continued)
-------
Section No. 3.9.1
Revision No. 0
Date January 4, 1982
Page 20 of 20
o
TABLE 1.1 (continued)
Apparatus
Reagents
Di stilled water
Crushed ice
Stopcock grease
Calcium oxide
powder
Phenolphthalein
Sodium hy-
droxide
Sulfuric acid
Fluoride stan-
dard solution
Acceptance limits
Must conform to ASTM-
01193-74, Type 3
Acetone insoluble, and
heat stable silicon
grease
Reagent grade or cer^
tified ACS
0.1% in 1:1 ethanol-
water mixture; reagent
grade or certified ACS
NaOH pellet 5M NaOH
reagent grade or cer-
tified ACS
Concentrated, reagent
grade or certified ACS;
25% (v/v) reagent grade
or ACS
Reagent grade or ACS;
1 M concentration
Frequency and method
of measurements
Check each lot
Check frozen condition
Upon receipt, check
label for grade or
certification
As above
As above
As above
As above
As above
Action if
requirements
are not met
Replace or re-
turn to manu-
facturer
As above
As above
As above
As above
As above
As above
O
o
( F
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 1 of 25
2.0 CALIBRATION OF APPARATUS
Calibration of apparatus is one of the most important func-
tions in maintaining data quality. The detailed calibration
procedures in this section are designed for the equipment speci-
fied in Method 13B and described in the previous section. A
laboratory log book of all calibrations must be maintained.
Table 2.1 at the end of this section summarizes the quality
assurance activities for calibration.
2.1 Metering System
2.1.1 Wet Test Meter - A wet test meter with a capacity of 3.4
m /h (120 ft /h) will be needed to calibrate the dry gas me,ter.
Wet test meters are calibrated by the manufacturer to an accuracy
of +0.5%; the calibration must be checked initially upon receipt
and yearly thereafter. For large wet test meters (>3£/rev),
there is no convenient procedure for checking the calibration;
for this reason, several methods are suggested, and others may be
approved by the administrator.
The initial calibration • may be checked by any of the fol-
lowing methods:
1. Certification from the manufacturer that the wet test
meter is within +1% of true value at the wet test meter dis-
charge, so that only a leak check is needed.
2. 'Calibration by any primary air or liquid displacement
method that displaces at least one complete revolution of the
wet test meter.
3. Comparison against a smaller wet test meter that has
previously been calibrated by a primary air or liquid displace-
ment method (Section 3.5.2).
4. Comparison against a dry gas meter that has previously
been calibrated by a primary air or liquid displacement method.
The calibration of the test meter should be checked annual-
ly. This yearly calibration check can be made by the same method
-------
Section -No. 3.9.2
Revision No. 0
Date January 4, 1982 f\
Page 2 of 25 V_y
as that of the original calibration; however, the comparison pro-
cedure need not be recalibrated if the check is within +1% of the
true value; if not within ±1%, either the comparison procedure or
the wet test meter must be recalibrated against a primary air or
liquid displacement method.
2.1.2 Sample Meter System - The sample meter system—consisting
of the pump, vacuum gauge, valves, orifice meter, and dry gas
meter—should be initially calibrated by stringent laboratory
procedures before it is used in the field. After initial accept-
ance, the calibration should be rechecked after each field test
series. The recheck procedure can be used by the tester often
and with little time and effort to ensure that calibration has
not changed. When the quick check indicates that the calibration
factor has changed, the tester must again use the complete labo-
ratory procedure to obtain a new calibration factor. ' After (j
recalibration, the metered sample volume must be multiplied by ^-^
either the initial or the recalibrated calibration factor—that
is, the one that yields the lower gas volume for each test run.
Before initial calibration, a leak check is recommended, but
it is not mandatory. Both positive (pressure) and negative
(vacuum) leak checks should be performed. Following is a pres-
sure leak-check procedure for checking the metering system from
the quick disconnect inlet to the orifice outlet and for checking
the orifice-inclined manometer:
1. Disconnect the orifice meter line from the downstream
orifice pressure tap (the one closest to the exhaust of the
orifice), and plug this tap (Figure 2.1).
2. Vent the negative side of the inclined manometer to the
atmosphere. If the manometer is equipped with a three-way valve,
merely turn the valve that is on the negative side of the
orifice-inclined manometer to the vent position. -^
3. Place a one-hole rubber stopper with a tube through ( j
its hole into the exit of the orifice, and connect a piece of
rubber or plastic tubing to the tube, as shown in Figure 2.1.
-------
RUBBER
TUBING
RUBBER ORIFICE
STOPPER
BLOW INTO TUBING UNTIL
MANOMETER READS 127 to
178 mm ( 5 TO 7 in.)
MAIN VALVE
CLOSED
ORIFICE/
MANOMETER
Figure 2.1. Positive leak check of metering system.
B> P> (T> fl>
iQ rt < O
fl> fO H-rt
w H-
H-O
os P3
O •
vo
CO
ro
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 4 of 25
4. Open the positive side of the manometer to the "read-
ing" position; if the manometer is equipped with a three-way
valve, this will be the line position.
5. Plug the inlet to the vacuum pump. If a quick discon-
nect with a leak-free check valve is used on the control module,
the inlet will not have to be plugged.
6. Open the main valve and the bypass valve.
7. Blow into the tubing connected to the end of the ori-
fice until a pressure of 127 to 178 mm (5 to 7 in.) H20 has built
up in the system.
8. Plug or crimp the tubing to maintain this pressure.
9. Observe the pressure reading for 1 min. No noticeable
movement in the manometer fluid level should occur. A bubbling-
type leak-check solution may aid in locating any leak in the
meter box. C)
After the metering system is determined to be leak free by
the positive leak-check procedure, check the vacuum system to and
including the pump.
1. Plug the air inlet to the meter box. If a quick dis-
connect with a leak-free stopper system is on the meter box, the
inlet will not have to be plugged.
2. Turn the pump on and pull a vacuum within 7.5 cm (3
in.) Hg of absolute zero.
3. Observe the dry gas meter. If leakage is > 0.00015
m3/min (0.005 fts/min), find and minimize the leak(s) until the
above specifications are satisfied.
For metering systems with diaphragm pumps, the leak-check
procedures above will not detect leakages within the pump; the
following procedure is suggested:
1. Make a 10-min calibration run at 0.00057 ms/min (0.02
ft3/min). ^^^
2. At the end of the run, find the difference between the ( J
measured wet test meter and the dry gas meter volumes, and divide
the difference by 10 to get the leak rate. The leak rate should
not exceed 0.00057 ms/min (0.02 ft3/min). •
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 5 of 25
2.1.2.1 Initial calibration - The dry gas meter and the orifice
meter can be calibrated simultaneously, and both should be cali-
brated when first purchased and any time the posttest check
yields a Y outside the range of the calibration factor Y +0.05Y.
Use a calibrated wet test meter (properly sized, with +1%
accuracy) to calibrate both the dry gas meter and the orifice
meter (Figure 2.2) in the following manner:
1. Leak check the metering system (Subsection 2.1.2), and
eliminate any leaks before proceeding.
2 . Connect the air outlet of the wet test meter to the
needle valve at the inlet side of the meter box (Figure 2.2).
3. Run the pump for 15 min with the orifice meter differ-
ential (AH) set at 12.7 mm (0.5 in.) H2O to allow the pump to
warm up and to permit the interior surface of the wet test meter
to be wetted.
4. Adjust the needle valve so that the vacuum gauge on the
meter box is between 50 and 100 mm (2 to 4 in. ) Hg during cali-
bration.
5. Record the required data on Figure 2.3A or 2.3B, using
sample volumes as shown.
6. Calculate Y. for each of the six runs, using the equa-
tion in Figure 2.3A or 2.3B, and 'record the results on the form
in the space provided.
7. Calculate the average Y (calibration factor) for the
six runs, using the following equation:
v
* ~
Record the average in the space provided on Figure 2.3A or
2.3B.
8. Clean, adjust, and recalibrate, or reject the dry
gas meter if one or more values are outside the interval Y
10.02Y; otherwise, the average Y is acceptable and should be
used for future checks and test runs .
-------
ORIFICE
X__XVCTT
IW
ORIFICE
AIR IKLET
LEVEL ADJUST
Figure 2.2. Sample meter system calibration setup,
HATER
LEVEL
GAUGE
HATER OUT
hJ t> Jd W
Q) P) 0 Q
iQ rf < O
0) a t->- rt
r. W H-
<*£H-O
r; o P
i^ *-* ^
(a
o
10
vo
oo
to
O
o
o
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 7 of 25
Date
S&
-
Barometric pressure, P, = c>?9,
Meter box number f=-H-
in. Hg Calibrated by
Orifice
manometer
setting
(AH),
in. H20
0.5
1.0
1:5
2.0
- 3.0
4.0
Gas volume
Wet test
meter
<'. + 46«
^ f£9, U V-") (S'<4 9^>
•T. / »)- W9 x 67^ /6"5A S")
.... 0.0317 AH (^w + 46°') 6
^i - Pb (t, + 460) [ Vw J
($>o3n)C<3~S^r?fr3j. S")^/p 75- ¥
4?3. (a*^)C .^^^ 1 5 -J
alf there is only one thermometer on the dry gas meter, record the temperature
under t,.
d
Figure 2.3A. Dry gas meter calibration data (English units).
(front side)
-------
Nomenclature :
V = Gas volume passing through the wet test meter, ft3.
w
V, = Gas volume passing through the dry gas meter, ft3.
t "= Temperature of the gas in the wet test meter, °F.
W
t, = Temperature of the inlet gas of the dry gas meter, °F.
t, = Temperature of the outlet gas of the dry gas meter, °F.
o
t, = Average temperature of gas in dry gas meter, obtained by average t, and
t °F i
CD n-rt
CD Q H- O
6 = Time for each calibration run, min. o S § 3
Hi C 25
P, = Barometric pressure, in. Hg.
Figure 2.3A. Dry gas meter calibration data (English units). (backside)
u>
o •
00
Jvj
ooo
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 9 of 25
Barometric pressure, P, =
Meter box number "p-hi -
mm Hg Calibrated by
Orifice
manometer
setting
(AH),
mm HgO
10
25
40
50
75
100
Gas volume
Wet test
meter
'/ C«29/ y
s only one thermometer on the d
,, 0.00117 AH ^w + 273^ 6
•** - Pb (td + 273) [ Vw J
(&.OO/1'7) (/(-^i f~C.?£> £ /O'£>«a}~
C73(i)C2lT) 1 OtJS'Z _J
ry gas meter, record the temperature
Figure 2.3B. Dry gas meter calibration data (metric units).
(front side)
-------
and
Nomenclature:
V = Gas volume passing through the wet test meter, m3.
w
V, = Gas volume passing through the dry gas meter, m3.
t = Temperature of the gas in the wet test meter, °C.
Vr
t, = Temperature of the inlet gas of the dry gas meter, °C.
t, = Temperature of the outlet gas of the dry gas meter, °C.
o
t, = Average temperature of gas in dry gas meter, obtained by average of t,
AH = Pressure differential across orifice, mm H2O.
Y. = Ratio of accuracy of wet test meter to dry gas meter for each run; tolerance Y. =
Y+0.02 Y.
Y = Average ratio of accuracy of wet test meter to dry gas meter for all six runs.
AH@. = Orifice pressure differential at each flow rate that gives 0.021 m3 of air at standard
conditions for each calibration run, mm H2O; tolerance AH@- = AH@±3.8 mm H2O
(recommended).
hj o to cr.
AH@ = Average orifice pressure differential that gives 0.021 m3 of air at standard con- ^ j»«| flj
ditions for all six runs, mm H2O; tolerance AH@ = 46.74 +6.3 mm H2O (recommended) CD ro n-rt
— W H
M C_| H- O
6 = Time of each calibration run, min. ° Ej g 3
o a a
P. = Barometric pressure, mm Hg. H, PJ z p
OJ
o •
vo
Figure 2.3B Dry gas meter calibration data (metric units). (backside) £
00 '
M
ooo
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 11 of 25
9. Calculate AH@^ for each of the six runs, using the
equation in Figure 2.3A or 2.3B, and record the results on the
form in the space provided.
10. Calculate the average AH@ for the six runs, using the
following equation:
AH@- + AH@ + AH@, + AH@A + 'AH@C + AH®.
AH@ = ± 2 3^ 4 5 6 .
D
Record the average in the space provided on Figure 2.3A or 2.3B.
11. Adjust the orifice meter or reject it if AH@. varies by
more than ±3.8 mm (0.15 in.) H20 over the range of 10 to 100 mm
(0.4 to 4.0 in.) H2O; otherwise, the average AH@ is acceptable
and should be used for subsequent test runs.
2.1.2.2 Posttest .calibration check - After each field test
series, conduct a calibration check of the metering system (Sub-
section 2.1.2) except for the following variations:
1. Three calibration runs at a single intermediate orifice
meter setting may be used with the vacuum set at the maximum
value reached during the test series. The single intermediate
orifice meter setting should be based on the previous field test.
To adjust the vacuum, insert a valve between the wet test meter
and the inlet of the metering system.
2. If a temperature-compensating dry gas meter was used,
the calibration temperature for the dry gas meter must be within
+6°C (10.8°F) of the average meter temperature during the test
series.
3. Use Figure 2.4A or 2.4B to record the required data.
If the recalibration factor Y deviates by <5% from the
initial Y (determined in Subsection 2.1.2), the dry gas meter
volumes recorded during the test series are acceptable; if Y
deviates by >5%, recalibrate the metering system (Subsection
2.1.2), and use the coefficient (initial or recalibrated) that
yields the lower gas volume for each test run.
Alternate procedures—for example, using the orifice meter
coefficients—may be used, subject to the approval of the admini-
strator.
-------
Test numbers /)<6 /~3
Barometric pressure, P
Date 5~/3-$O Meter box number
~ 7
Plant flcme.
in. Hg Dry gas meter number
Pretest Y
p.
Orifice
manometer
setting,
(AH),
in. H20
"AW
Gas volume
Wet test
meter
(vw),
ft3
10
10
10
Dry gas
meter
,
°F
7«
Dry gas meter
Inlet
°F
^3
Outlet
(td),
o
°F
75"
Average
°F
79
Time
(9),
nin
7*35-
Vacuum
setting,
in. Hg
3-o
Yi
0-95-7
Y.
V P. (t. + 460)
w b x d '
VIP + ... ._ iff i /inni
/o (*l& • 7»OC*'5~.3<50
/0.*A3(Z9.7*+-^)(S3*
Y =
V
If there is only one thermometer oh the dry gas meter, record the temperature under t..
V = Gas volume passing through the wet test meter, ft3.
w
V. = Gas volume passing through the dry gas tneter, ft3.
t = Temperature of the gas in the wet test meter, °F.
W
t. = Temperature of the inlet gas of the dry gas meter, °F.
t. = Temperature of the outlet gas of the dry gas meter, °F.
o
t. = Average temperature of the gas in the dry gas meter, obtained by the average of
AH = Pressure differential across orifice, in. H20.
Y. = Ratio of accuracy of wet test neter to dry gas meter for each run.
Y = Average ratio of accuracy of wet test meter to dry gas meter for all three runs;
tolerance = pretest Y +0.05Y
P. = Barometric pressure, in. Hg.
8 = Time of calibration run, nin.
Figure 2.4A. Posttest dry gas meter calibration data form (English units).
D» a-
and t. , °F. ° " K"
QO M Ci H-
u ro p o
n>
H|P»
in
U>
•
vo
vO
OJ
O
o
o
-------
lest numbers
1-3 Date S-/3-SO Meter box number
Barometric pressure, P. =
mm Hg Dry gas meter number /r/y?- 7
Plant Acme, fkcoer
Pretest Y
Orifice
manometer
setting,
(AH),
mm H20
3&>
r Gas volume
Wet test
meter
(Vw>,
m3
0.30
0.30
0.30
Dry gas
meter
(vd),
m3
24. / 744
l
°C
33.5"
Outlet
(td),
o
°C
AL5"
Average
(td),*
°C
£3.5"
Time
(e),
min
I3.S-0
Vacuum
setting,
mm Hg
75"
Y.
o. <&&>
Yi
Vw Pb (td * 273)
V /P. + AH Vt + 273\
dlb Ti^Aw )
0.30 C73o)Ca5'.5-y)
o.3or3.C733.6,Ka<0
Y =
If there is only one thermometer on the dry gas meter, record the temperature under td'
V = Gas volume passing through the wet test meter, m3.
w
Vd = Gas volume passing through the dry gas meter, m3.
t = Temperature of the gas in the wet test meter, °C.
W
t. = Temperature of the inlet gas of the dry gas meter, °C.
di
td = Temperature of the outlet gas of the dry gas meter, °C.
o
td = Average temperature of the gas in the dry gas meter, obtained by the average of td and td , °C.
AH = Pressure differential across orifice, in. H20.
Y. = Ratio of accuracy of wet test meter to dry gas meter for each run.
Y = Average ratio of accuracy of wet test meter to dry gas meter for all three runs;
tolerance = pretest Y +0.05Y
Pb = Barometric pressure, in. Hg.
9 = Time of calibration run, min.
Figure 2.4B. Posttest meter calibration data form (metric units).
^ t2 jd OT
CD ft> H-rt
en H-
H ^4 H- O
O C 2
HI P a o
O •
Ul
(A)
to
CO
ro
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 14 of 25
o
o
2.2 Temperature Gauges
2.2.1 Impinger Thermometer - The thermometer used to measure
temperature of the gas leaving the impinger train should initial-
ly be compared with a mercury-in-glass thermometer which meets
ASTM E-l No. 3C or 3F specifications. The procedure is as
follows:
1. Place both the reference thermometer and the test
thermometer in an ice bath, and compare readings after they
stabilize.
2. Remove the thermometers from the bath, and allow both
to come to room temperature, compare readings after they stabi-
lize.
3. Accept the test thermometer if both of its readings
agree within 1°C (2°F) of the reference thermometer reading. If
the difference is greater than ±1°C (2°F), either adjust and
recalibrate it until agreement is obtained, or reject it.
2.2.2 Dry Gas Thermometers - The thermometers used to measure
the metered gas sample temperature should initially be compared
with a mercury-in-glass thermometer, using a similar procedure.
1. Place a dial type or an equivalent thermometer and a
mercury-in-glass thermometer in a hot water bath, 40° to 50°C
(105° to 122°F); compare the readings after the bath stabilizes.
2. Allow both thermometers to come to room temperature and
compare readings after they stabilize.
3. Accept the dial type or equivalent thermometer if the
values agree within 3°C(5.4°F) at both points or if the temper-
ature differentials at both points are within ±3°C(5.4°F); tape
the temperature differential to the thermometer/ and record them
on the pretest sampling check form (Figure 3.1 of Section 3.9.3).
4. Before each field trip, compare the reading of the
mercury-in-glass thermometer at room temperature with that of the
meter system thermometer; the values or the corrected values [)
should agree within ±6°C (10.8°F) of one another, or the meter \*-S
thermometer should be replaced or recalibrated. Record any
correction/factors on Figure 3.1 or on^a similar form. / Ac/
M'1
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
i Page 15 of 25
2.2.3 Stack Temperature Sensor - The stack temperature sensor
should be calibrated upon receipt or checked before field use.
Each sensor should be uniquely marked for identification. The
calibration should be performed at three points and then extra-
polated over the temperature range anticipated during actual
sampling. For the three-point calibration, a reference mercury-
in-glass thermometer should be used. The following procedure is
recommended for calibrating stack temperature sensors (thermo-
couples and thermometers) for field use,
1. For the ice point calibration, form a slush from
crushed ice and water (preferably deionized distilled) in an
insulated vessel such as a Dewar flask; being sure that the
sensor does not touch the sides of the flask, insert the stack
temperature sensor into the slush to a depth of at least 2 in.
Wait 1 min to achieve thermal equilibrium, and record the readout
on the potentiometer. Obtain three readings taken in 1-min
intervals. Note; Longer times may be required to attain thermal
equilibrium with thick-sheathed thermocouples.
2. Fill a large Pyrex beaker with water to a depth >4 in.
Place several boiling chips in the water, and bring the water to
a full boil using a hot plate as the heat source. Insert the
stack temperature sensor(s) in the boiling water to a depth of at
least 2 in., taking care not to touch the sides or bottom of the
beaker.
Alongside the sensor(s), an. ASTM reference thermometer
should be placed. After 3 min, both instruments will attain
thermal equilibrium. Simultaneously record temperatures from the
ASTM reference thermometer and the stack temperature sensor three
times at 1-min intervals.
If the entire length of the mercury column in the thermom-
eter cannot be immersed, a temperature correction will be re-
quired to give the correct reference temperature.
3. For a thermocouple, repeat Step 2 with a liquid (e.g.,
cooking -oil) that has a boiling point 150° to 250°C (300° to
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 16 of 25
500°F), and record all data on Figure 2.5. For thermometers
other than thermocouples, either repeat Step 2 with a liquid that
boils at the maximum temperature that the thermometer is to be
used or place the stack thermometer and the reference thermometer
in a furnace or other device to attain the desired temperature.
Note; if the thermometer is to be used at temperatures higher
than the reference thermometer will record, calibrate the stack
thermometer with a thermocouple previously calibrated by the
above procedure.
4. If the absolute temperature of the reference thermom-
eter and the thermocouple(s) agree within ±1.5% at each of the
three calibration points, either plot the data on' linear graph
paper and draw the best-fit line between the points or calculate
the linear equation using the method of least-squares. For the
thermocouple, the data may be extrapolated above and below the C)
calibration points to cover the manufacturer's suggested range," ^-""^
for the portion of the plot (or equation) that agrees within
1.5% of the absolute reference temperature, no correction is
needed, but for all other portions that do not agree within
±1.5%, use the plot (or equation) to correct the data.
If the absolute temperatures of the reference thermometer
and stack temperature sensor (other than the thermocouple) agree
within ±1.5% at each of the three points, the thermometer may be
used for testing without applying any correction factor over the
range of calibration points, but the data cannot be extrapolated
outside the calibration points.
2.3 Probe Heater
The probe heating system should be calibrated before field
use according to the procedure in APTD-0576.*1 Probes constructed
according to APTD-05813 need not be calibrated if the curves of
APTD-05764 are used. ^^^
2.4 Barometer f j
The field barometer should be adjusted initially and before
each test series to agree within ±2.5 mm (0.1 in-.) Hg of the
mercury-in-glass barometer reading or with the value reported by ,
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 17 of 25
Date
Thermocouple number
Ambient temperature
Calibrator
C Barometric pressure
J. £ 7
in. Hg
Reference: mercury-in-glass
other
Reference
point
number
0°
100'
Source3
(specify)
•/C£ «>*TtA,
Goi/i'**} Utfi'T£&
DOI/I'A/Q C It &->>/4
Oil
Reference
thermometer
temperature ,
°C
r -
/£>/. 5~'
Thermocouple
potentiometer
temperature,
°C
/'
/*''
Temperature.
difference,
•
OJ1-
Type of calibration system used.
3f(ref temp. °C + 273) - (test thermom temp, °C + 273)1
L ref temp, "C + 273 \
Figure 2.5 Stack temperature sensor calibration data form.
-------
o
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 18 of 25
a nearby National Weather Service station and corrected for
elevation. Correction for the elevation difference between the
station and the sampling point should be applied at a rate of
-2.5 mm Hg/30 m (0.1 in. Hg/100 ft). Record the results on the
pretest sampling check form (Figure 3.1).
2.5 Probe Nozzle
Probe nozzles should be calibrated initially before use in
the field.
1. Use a micrometer to measure the ID of the nozzle to the
nearest 0.025 mm (0.001 in.).
2. Make three measurements using different diameters each
time.
3. Average the three measurements. The difference between
the high and the low numbers should not be <_0.1 mm (0.004 in.).
4. Label each nozzle permanently and uniquely for Identi- x—x
fication. \^/
5. Record the data on Figure 2.6, the nozzle calibration
form. If nozzles become nicked, dented, or corroded, reshape,
sharpen, and recalibrate before use.
2.6 Pitot Tube
The Type S pitot tube assembly should be calibrated using
the procedure in Section'3.1.2, Method 2.
2.7 Trip Balance
The trip balance should be calibrated initially by using
Class-S standard weights, and it should agree within ±0.5 g of
the standard weight. Adjust or return the balance to the manu-
facturer if limits are not met.
2.8 Fluoride Electrode
The fluoride (F~) electrode should be calibrated daily, and
checked hourly against serial dilutions of the 0.1M fluoride
standard solution. Use the following procedure to prepare and to
measure the concentration of the dilutions. f j
1. Pipette 10 ml of 0.1M NaF into a 100-ml volumetric
flask, and dilute to the mark with distilled water to get
0.01M NaF;,'Cdilute 10 ml of the 0.01M solution to make a 0.001M ; :' '
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 19 of 25
Date
1 tO
Calibrated by
Nozzle
identification
number
37
Nozzle Diameter3
jnar?ii.)
/< "^ *-T" /
U- ^° /
D2,
jffin (in. )
D3,
mm (in. )
AD,b
mm (in. )
D c
avg
where:
"D
1,2,3,
AD =
three different nozzles diameters, mm (in.); each
diameter must be within (0.025 mm) 0.001 in.
maximum difference between any two diameters, mm (in.),
AD £(0.10 mm) 0.004 in.
avg
= average of D,, D2/ and
Figure 2.6 Nozzle calibration data form.
-------
Section No. 3.9.2
Revision No. -<
Date January 4, 1982
Page 20 of 25
solution; continue in the same manner to get the 0.001M and the
0.00001M dilutions.
2. Pipette 50 ml of each NaF dilution into separate
beakers.
3. Add 50 ml of TISAB to each beaker.
4. Immerse the electrode into the most dilute standard
solution, and measure the developed potential while stirring the
solution with a magnetic stirrer. Note; Avoid stirring the
solution before immersing the electrode because entrapped air
around the crystal can cause erroneous readings or needle fluc-
tuations .
5. Keep the electrodes immersed in the solution 3 min in
order for it to stabilize before taking a final positive milli-
volt reading.
6. Record the reading on the laboratory form, Figure 2.7,
and remove the electrode from the sample.
7. Soak . the electrode for 30 s in distilled water, and
then blot it dry.
8. Plot the millivolt value on the linear axis of semilog
graph paper and plot the known concentrations of fluoride stan-
dards on the log axis as shown in Figure 2.8. Note ; Plot the
nominal value for concentrations of the standards on the log
axis; for example, when 50 ml of the 0.01M standard is diluted
with 50 ml TISAB, the concentration is plotted as 0.01M. Measure
the most dilute standard first and the most concentrated standard
last, as shown on Figure 2.8, to get a straight-line calibration
curve with nominal concentrations of 0.00001, 0.0001, 0.001,
0.01, and 0.1M NaF. To obtain the required precision, use 4 or 5
cycle semilog paper similar to that in Figure 2.8.
To check the accuracy of the calibration curve, prepare and
measure a control sample (Section 3.9.5). Prepare fresh stan-
dardizing solutions of <_0.01M NaF daily, and store the solutions
in polyethylene or polypropylene contains.
The fluoride electrode should be .checked periodically after
repeated use for responsiveness and sensitivity. Compare the
o
o
(\j
\r
-------
section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 21 of 25
LABORATORY WORKSHEET
Date standards prepared _
Temperature of standards
, S* C.
Date J- S~ SO
Electrode number
Standard number
/
£
3
4
5"
£
Control Sample
Concentration (M)
0.000001
0.00001
0.0001
0.001
0.01
0.1
o.oor
Electrode potential (mV)
—
300
A<*7
£01
1*1
10
ItU
Note: The concentration of the control sample determined from the calibration
curve must be between 0.002M and 0.01M.
Signature of analyst
Signature of reviewer
Figure 2.7. Fluoride calibration data form. (Method 13B)
i 11
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 22 of 25
o
50
100
150
I 200
250
Date .
Sample temp
Analyst T" Loco n
Reviewer k) M/ t£f/u> / /
300
10
Results
Mo1 a ri ty
0.00001M
0.0001M
0.001M
0.01M
0.1M
Control
sample
-4
-3
10 10
FLUORIDE MOLARITY (M)
10
-2
nW
3 o
10
-1
O
Figure 2.8. Fluoride calibration curve, Method 13B.
O
< K
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 23 of 25
electrode responses or millivolt readings of samples or standards
of the same fluoride concentration. For equal concentrations of
fluoride, the electrode response should remain stable with each
analysis, if not, repair or replace the electrode.
Certain specific-ion meters designed specifically for fluoride
electrode use give direct readouts of F~ concentrations. These
meters may be used over narrow concentration ranges. Calibrate
the meter according to manufacturer's directions.
-------
Section No. 3.9.2
Revision No. .0
Date January 4, 1982
Page 24 of 25
o
TABLE 2.1. ACTIVITY MATRIX FOR CALIBRATION OF EQUIPMENT
Apparatus
Wet test meter
Dry gas meter
Thermometers
Barometer
Probe nozzle
(continued)
Acceptance limits
Capacity of >3.4 m3/h
(120 fts/h)f accuracy
within ±1.0%
Y. = Y ± 0.02 Y at
flow rate of 0.02 -
0.03 mVmin (0.7 -
1.1 ftVmin)
Impinger thermometer
+1°C (2°F); dry gas
meter thermometer
+3°C (5.4°F) over
applicable range
+2.5 mm (0.1 in.) Hg of
mercury-in-glass barom-
eter
Average three ID mea-
surements of nozzle;
difference between high
and low <0.1 mm
(0.004 in.)
Frequency and method
of measurement
Calibrate initially
and yearly by liquid
displacement
Calibrate with wet
test meter initially
to agree within Y ±
0.02 Y and when post-
test check is not
within Y ± 0.05 Y
Calibrate each ini-
tially against a
mercury-in-glass
thermometer; before
field trip compare
each with mercury-
in-glass thermometer
Calibrate initially
vs mercury-in-glass
barometer; check
before and after
each field test
Use a micrometer to
measure to near-
est 0.025 mm (0.001
in.)
Action if
requirements
are not met
Adjust to
meet specifi-
cations, or
return to
manufacturer
Repair or re-
place, and
then recali-
brate
Adjust, de-
termine a
constant cor-
rection fac-
tor, or re-
ject
Adjust to
agree with
certified
barometer
Recalibrate,
reshape, and
sharpen when
nozzle be-
comes nicked,
dented, or
corroded
O
o
,J
-------
Section No. 3.9.2
Revision No. 0
Date January 4, 1982
Page 25 of 25
Table 2.1 (continued)
Apparatus
Stack tempera-
ture sensor
Trip balance
Pi tot tube
Fluoride
electrode
Acceptance limits
±1.5% of average stack
temperature, °R
Standard Class-S
weights within ±0.5 g
of stated value
Type S; initially
calibrated according to
Section 3.1, Meth 2;
tube tips undamaged
Calibration curve plot-
ted with F standard
solutions of 0.1M,
0.01M, 0.001M,
0.0001M, and 0.00001M
and corresponding mV
reading on semi log
graph paper; stable
electrode response
Frequency and method
of measurement
Calibrate initially;
check after each
field test
Verify calibration
when first purchased,
any time moved or
subject to .rough
handling, and during
routine operations
when not within
± 0.5 g
Visually check
before each field
test
Calibrate with each
use and every hour
of continuous use;
check response
stability of elec-
trode after re-
peated use
Action if
requirements
are not met
Adjust or
reject
Have the
manufacturer
recalibrate
or adjust
Repair or
replace
Repair or
replace
-------
o
o
o
-------
Section No. 3.9.3
Revision No. 0
Date January 4, 1982
Page 1 of 6
3.0 PRESAMPLING OPERATIONS
The quality assurance activities for presampling operations
are summarized in Table 3.1 at the end of this section. See
Sections 3.0.1 and 3.0.2, of this Handbook for general informa-
tion on preliminary site visits. See Section 3.4.3 (Method 5)
for more detailed and specific information of presampling opera-
tions for sampling equipment similar to the Method 13 equipment.
3.1 Apparatus Check and Calibration
Pretest checks must be made on most of the sampling apparat-
us. Figure 3.1 should be used to record data on the pretest
calibration checks. Figure 3.2 in Section 3.4 of this Handbook
is recommended to aid the tester in preparing an equipment check-
list, status form, and packing list.
3.1.1 Sampling Train - A schematic of the EPA Method 13 sampling
train (Figure 1.1) should be used for assembling the components
and for checking for compliance (specifications in the Reference
Method, Section 3.9.10).
3.1.2 Probe and Noz'zle - Clean the probe and the nozzle inter-
nally by brushing first with tap water, then with deionized dis-
tilled water, and finally with acetone; allow both to dry in the
air. The probe should be sealed at .the inlet or tip; should be
checked for leaks at a vacuum of 380 mm (15 in.) Hg; and should
be leak free under these conditions. The probe liner, in extreme
cases, can be cleaned with stronger reagents; in either case, the
objective is to leave the liner free from contaminants. Check
the probe's heating system to see that it is operating properly
and that it prevents moisture condensation.
3.1.3 Impingers, Filter Holders, and Glass Connectors - All
glassware should be cleaned first with detergent and tap water
and then with deionized distilled water. All glassware should be
visually inspected for cracks or breakage and then repaired or
discarded if defective. If a filter is to be used between the
probe and the first impinger be sure that an acceptable stainless
steel mesh filter support is packed; glass frit supports are not
acceptable. / V 7
-------
Section No. 3.9.3
Revision No. 0
Date January 4, 1982
Page 2 of 6
Date $~-//-S>O Calibrated by
Meter box number F/3- / AH@ /. S 7
Dry Gas Meter*
Pretest calibration factor Y /. O/3 (within ±2% of the
average factor for each calibration run).
Impinger Thermometer
Was a pretest temperature correction used? yes *x"_ no
If yes, temperature correction r (within ±1°C (2°F)
of reference values for calibration and within ±2°C (4°F) of
reference value for calibration check)
Dry Gas Meter Thermometers
Was a pretest temperature correction made? yes *^ no
If yes, temperature correction (within ±3°C (5.4°F) of
reference value for calibration and within 6°C (10.8°F) of
reference values for calibration check)
Stack Temperature Sensor*
Was a stack temperature sensor calibrated against a reference
thermometer? ^" yes __^ ^__ no
If yes, give temperature range with which the readings agreed
within ±1.5% of the reference values 3oo to 36>o K
Barometer
Was the pretest field barometer reading correct? t^yes no
(within ±2.5 mm (0.1 in.) Hg of the mercury-in-glass barometer)
Nozzle*
Was the nozzle calibrated to the nearest 0.025 mm (0.001 in.)?
yes no
*Most significant items/parameters to be checked.
Figure 3.1. Pretest sampling checks
o
O
'i)C
-------
Section No. 3.9.3
Revision No. 0
Date January 4, 1982
Page 3 of 6
3.1.4 Pump - The vacuum pump should be serviced as recommended
by the manufacturer, or every 3 mo, or upon erratic behavior
(nonuniform or insufficient pumping action). Check the oiler
jars, if used, every 10 tests.
3.1.5 Dry Gas Meter - A dry gas meter calibration check should
be made using the procedure in Section 3.4.2.
3.1.6 Silica Gel - Either dry the used silica gel at 175°C
(347°F) for at least 2 h or use fresh silica gel. Weigh several
200- to 300-g portions in airtight containers to the nearest
0.5 g if the moisture content is to be determined. Record the
total weight (silica gel plus container) for each container.
3.1.7 Thermometers - The thermometers should be compared to a
mercury-in-glass reference thermometer at ambient temperature.
3.1.8 Barometer - The field barometer should be compared before
each field trip with a mercury -in -glass barometer or with a
weather station reading after making an elevation correction.
3-2 Reagents and Equipment
3.2.1 Filters - Check the filters visually against light for
irregularities, flaws, and pinhole leaks. Determine the F-blank
value by analyzing three filters chosen from each lot (Section
3.9.5); if the value is <_0.015 mg F/cm2 they are acceptable for
field use.
3.2.2 Water - 100 ml of deionized distilled water is needed for
each of the first two impingers and for sample recovery.
3.2.3 Ice - Crushed ice is needed to keep the gas that exits
into the last impinger below 21°C (70°F).
3.2.4 Stopcock grease - Silicone grease that is acetone insol-
uble and heat stable may be used sparingly at each connection
point of the sampling train to prevent gas leaks; but this is not
necessary if screw-on connectors with Teflon (or similar) sleeves
are used.
3.3 Equipment Packing
The accessibility, condition, and functioning of measurement
devices in the field depend on careful packing and on careful
-------
Section No. 3.9.3
Revision No. 0
Date January 4, 1982 C J
Page 4 of 6
movement on site. Equipment should be packed to withstand severe
treatment during shipping and field handling operations. One
major consideration in shipping cases is the construction mate-
rials. The following containers are suggested, but are not
mandatory.
3.3.1 Probe - Seal the inlet and outlet of the probe to protect
the probe from breakage. Pack the probe inside a container that
is lined with polyethylene or other suitable material that is
rigid enough to prevent bending or twisting during shipping and
handling; an ideal container is a wooden case (or the equivalent)
with a separate compartment lined with foam material for each
probe, and should have handles or eye-hooks .that can withstand
hoisting.
3.3.2 Impingers, Connectors, and Assorted Glassware - All im-
pingers and glassware should be packed in rigid containers and ( J
protected by polyethylene or other suitable material. Individual
compartments will help to organize and protect each piece of
glassware.
3.3.3 Volumetric Glassware - A sturdy case lined with foam
material is suggested for drying tubes and assorted volumetric
glassware.
3.3.4 Meter Box - The meter box—which contains the manometers,
orifice meter, vacuum gauge, pump, dry gas .meter/ and thermom-
eters—should be packed in a shipping container unless its hous-
ing is sufficiently protective for the components during travel.
Additional pump oil should be packed if oil is required. It is
advisable to carry a spare meter box in case of failure.
3.3.5 Wash Bottles and Storage Containers - Storage containers
and miscellaneous glassware should be packed in "a rigid foam-
lined container.
o
1'
-------
Section No. 3.9.3
Revision No. 0
Date January 4, 1982
Page 5 of 6
TABLE 3.1 ACTIVITY MATRIX FOR PRESAMPLING OPERATIONS
Apparatus
Sampling train
probe and
nozzle
Impingers,
filter
holders, and
glass con-
nectors
Pump
Dry gas meter
Acceptance limits
1. Probe, nozzle, and
liner free of contami-
nants; constructed of
borosilicate glass,
quartz, or equivalent;
metal liner must be
approved by admini-
strator
2. Probe leak free
at 380 mm (15 in.) Hg
3. Probe heating
system prevents mois-
ture condensation
Clean; free of breaks,
cracks, leaks, etc.
Sampling rate of 0.02-
0.03 mVmin (0.7 to
1.1 fts/min) up to 380
mm (15 in.) Hg at pump
inlet
Clean; readings ±2% of
of average calibration
factor
Frequency and method
of measurements
1. Clean internally
by brushing with tap
water, deionized dis-
tilled water, and
acetone; air dry
before test
2. Check using pro-
cedures in Subsec 2.3
3. Check heating
system initially and
when moisture cannot
be prevented during
testing (Sec 3.4.1)
Clean with detergent,
tap water, and
deionized distilled
water
Service every 3 mo
or upon erratic be-
havior; check
oiler jars every 10
tests
Calibrate according
to Sec 3.4.2; check
for excess oil
Action if
requirements
are not met
1. Repeat
cleaning and
assembly pro-
cedures
2. Replace
3. Repair or
replace
Repair or
discard
Repair or re-
turn to manu-
facturer
As above
(continued)
v..
-------
Section No. 3.9.3
Revision No. 0
Date January 4, 1982
Page 6 of 6
Table 3.1 (continued)
o
Apparatus
Reagents and
Equipment
Filters
Water
Stopcock grease
Packing Equip-
ment for
Shipment
Probe
Impingers, con-
nectors, and
assorted
glassware
Pump
Meter box
Wash bottles
and storage
containers
Acceptance limits
No irregularities,
flaws, pinhole leaks;
<0.015 mgF/cm2
Deionized distilled
conforming to
ASTM-D1193-74, Type 3
Acetone insoluble;
heat stable
Rigid container lined
with polyethylene foam
Rigid container lined
with polyethylene foam
Sturdy case lined with
polyethylene foam ma-
terial if not part of
meter box
Meter box case and/or
additional material to
protect'train compon-
ents; pack spare meter
box
Rigid foam-lined con-
tainer
Frequency and method
of measurements
Visually check before
testing; check each
lot of filters for F
content
Run blank evapora-
tions before field
use to eliminate high
solids (only required
if impinger contents
to be analyzed)
Check label
receipt
upon
Prior to each ship-
ment
As above
As above
As above
As above
Action if
requirements
are not met
Replace
Redistill
replace
or
Replace
O
Repack
As above
As above
As above
As above
O
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 1 of 21
4.0 ON-SITE MEASUREMENTS
The on-site activities include transporting equipment to the
test site, unpacking and assembling the equipment, making duct
measurements, performing the velocity traverse, determining
molecular weights and stack gas moisture contents, sampling for
fluorides, and recording the data. Table 4.1 at the end of this
section summarizes the quality assurance activities for on-site
activities. Blank forms to be used in recording data are in Sec-
tion 3.4.12 for the convenience of the Handbook user.
4.1 Handling of Equipment
The most efficient means of transporting or moving the
equipment from ground level to the sampling site should be de-
cided during the preliminary site visit (or prior correspon-
dence). Care should be exercised to prevent damage to the test
equipment or injury to test personnel during the moving phase. A
"laboratory" area should be designated for assembling the sam-
pling train, placing the filter in the filter holder, charging
the impingers, recovering the sample, and documenting the re-
sults; this area should be clean and free of excessive drafts.
4.2 Sampling
The on-site sampling includes preliminary measurements and
setup, placing the filter in the filter holder, setting up the
sampling train, preparing the probe, checking for leaks along the
entire train, inserting the probe into the stack, sealing the
port, checking the temperature of the probe, sampling at desig-
nated points, and recording the data. A final leak check must
always be performed upon completion of the sampling.
4.2.1 Preliminary Site Measurements - These measurements are
needed for locating the pitot tube and the probe during the
sampling.
1. Be sure the site meets Method 1 specifications; if not,
due to duct configuration or other- reasons, have the site ap-
proved by the administrator.
'ft
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982 x—s
Page 2 of 21 I)
2. Be sure a 115-V, 30-A electrical supply is available
for operating the standard sampling train.
3. Either measure the stack and determine the minimum
number of traverse points by Method 1, or check the traverse
points determined during the preliminary site visit (Section
3.0).
4. Record all data on the traverse point location form
shown in Method 1. These measurements will be used to locate the
pitot tube and the sampling probe during preliminary measurements
and actual sampling.
The site must be accepted before a valid sample can be taken.
4.2.2 Stack Parameters - By the following preliminary measure-
ments, the user can set up the nomograph as outlined in
APTD-0576.4 An example nomograph data form is Figure 4.1. Using
the stack parameters obtained, the isokinetic sampling rate can
be set.
1. Check the sampling site for cylonic or nonparallel flow
(Method 1, Section 3.0).
2. Determine the stack pressure, temperature, and the ';
range of velocity heads encountered (Method 2).
3. Calculate the moisture content by using Method 4 or
its alternatives. If the source has been tested before or if a
good estimate of the moisture is available, this value can be
used to avoid calculations. If the stack gas is saturated with
moisture or has water droplets, the moisture content must be
determined by using stack gas temperature sensor (Method 4).
4. Calculate the dry molecular weight (M,) of the stack
gas (Method 2). If an integrated gas sample is required, follow
Method 3 procedures and take the gas sample simultaneously with
and for the same total length of time as the fluoride sample.
5. Record the data on the sampling and the analytical data
forms for molecular weight determinations located in Section 3.2, [)
Method 3. ^-^
Note; The condensate collected during the sampling can be used
in the final calculations of moisture content (Section 3.9.6). ,
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 3 of 21
Plant
Date
Sampling location
Calibrated pressure differential across
orifice, in. H2O
Average meter temperature (ambient + 20°F), °F
Percent moisture in gas stream by volume, %
Barometric pressure at meter, in. Hg
Static pressure in stack, in. Hg
(Pm±0.073 * stack gauge pressure, in. H2
Ratio of static pressure to meter pressure
Average stack temperature, °F
Average velocity head, in. H2O
Maximum velocity head, in. H2O
C factor
Calculated nozzle diameter, in
Actual nozzle diameter, in.
Reference Ap, in. H2O
m
avg
wo
avg
AP
avg
SO
a 1.
- O-OI
0-3
/.
0.38S
0. 37S
0-148
Figure 4.1. Nomograph data form (English units).
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 4 of 21
4.2.3 Sampling Rate - Method 13 sampling is performed isokinet-
3
ically like- Method 5, but the sampling rate must be <0.03 m /mm
3
(1.0 ft /min) during the test; the maximum AH will limit the rate
to <0.03 m3/min (1.0 ft3/min).
1. Select a nozzle size based on the range of velocity
heads, so that the nozzle size will not have to be changed to
maintain an isokinetic sampling rate.
2. Select a nozzle that will maintain the maximum sampling
3 3
rate at <0.03 m /min (1.0 ft /min) during the run.
3. Check the maximum AH, using the following equation:
1.09 P M AH@
Maximum AH <_ - = - Equation 4-1
m
where
Maximum AH = pressure differential across the orifice that
produces a flow of 1.0 ft /min, in. H20;
P = pressure of the dry gas meter, in. Hg;
M = molecular weight of the stack gas;
AH@ = pressure differential across the orifice that
produces a flow rate of 0.75 scfm, in. H20; and
. T = temperature of the meter, °R.
4. Install the selected nozzle using a Viton A 0-ring if
glass or stainless steel liners are used; install the nozzle on a
stainless steel liner by using a leak-free mechanical connection
(APTD-05764) or Teflon ferrules.
5. Mark the probe with heat resistant tape or with another
acceptable means to denote the proper distance to which it should
be inserted into the stack or duct at each sampling point.
6. Select a total sampling time that is greater than or
equal to the minimum total sampling time specified in the test /"""N
procedures for the specific industry so that — ^ — ^
a. The sampling time per traverse point is >2 min
(greater time interval may be specified by the administrator); .— y/(-
'
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 5 of 21
the number of minutes sampled at each point is either an integer
or an integer plus one-half minute to avoid timekeeping errors.
b. The sample volume corrected to standard conditions
exceeds the required minimum total gas sample volume (can be
based on an approximate average sampling rate). In some circum-
stances (e.g., batch cycles), it may be necessary to sample for
shorter times and to obtain smaller gas sample volumes; if so,
obtain the administrator's approval.
7. Record the data on the fluoride field data form (Figure
4.2).
4..2.4 Sampling Train Preparation - These steps are needed for
preparing the sampling train.
1. Keep all openings where contamination can occur covered
until just before assembly of the setup or before beginning the
sampling.
2. Place 100 ml of distilled water (a graduated cylinder
may be used) in each of the first two impingers.
3. Leave the third impinger empty.
4. Add 200-300 g of preweighed silica gel in the fourth
impinger and place the empty container in. a safe place for use
later in the sample recovery, and record the weight of the silica
gel and the container on the appropriate data form. If moisture
content is to be determined by impinger analysis, weigh each of
the first three impingers to the nearest 0.5 g, and record these
weights.
5. Use tweezers or clean disposable surgical gloves to
place the filter in the filter holder.
6. Be sure that the filter is properly centered and that
the gasket is properly placed to prevent the sample gas stream
from circumventing the filter.
4.2.5 Sampling Train Assemblage - The arrangement of .the sam-
pling train components is shown in Figure 1.1.
1. Apply if needed to avoid contamination a very light
coat of silicone grease, but only on the outside of all ground-
glass joints. ' /t'l//
-------
Plant
faicr
Locatio
Operator
Date .T~
~ fiO
Run number
Stack diam, xz
Meter calibration (Y) _
Pi tot tube (C ) f).f}4
Probe length p//? x/
Probe liner material ^
Probe heater setting ^
Ambient temperature
Sheet
Nozzle
of
identification
Nozzle diameter ft
Thermometer number
Final leak rate
7_
number
mm (in.
7/9,3
Sample box nuraber
Heter box number •:
Meter AH9
Barometric pressure (•.
Assumed moisture ^""
Static pressure (
C Factor /?O£,
-
cn^
£L
Q.39
&L.
£V_
< o
H-d-
tn H-
H-O
O J3
37
3.5-
3/3
61
•57
6/
-793
(Ft
O •
O •
\0
3.0
n.zn
/.
Total
Total
Hax
Figure 4.2. Fluoride field data form.
vo
oo
ro
O
o
o
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 7 of 21
2. Either place the filter immediately following the probe
or between the third and fourth impingers. Normally the filter
will be placed after the third impinger unless the filterable
particulate fluoride is to be measured. It is not necessary to
have filters in both positions. If in the front filter position,
use a 20-mesh stainless steel filter support with a silicone
outer seal.
3. Place crushed ice and water around the impingers.
4. Check the filter to be sure there are no tears.
5. Attach a temperature sensor to the metal sheath of the
sampling probe if it is not already an integral part of the
assembly, so that the sensor extends beyond the probe tip and
does not touch any metal; the sensor should be about 1.9 to
2.54 cm (0.75 to 1 in.) from the pitot tube and the nozzle to
avoid interference with the gas flow; other arrangements are
shown in Method 2.
4.2.6 Sampling Train Leak Checks - Leak checks are necessary to
ensure that the sample has not been biased low by dilution air.
The Reference Method (Section 3.9.10) specifies that leak checks
be performed at certain times as discussed below. Leakage rates
<4% of the average sampling rate or 0.00057 m3/min (0.02
ft3/min), whichever is less, are acceptable.
4.2.6.1 Pretest - (optional) If the tester opts to conduct the
pretest leak check, the following procedure should be used after
the sampling train has been assembled.
1. Set the filter heating system at the desired operating
temperature.
2. Allow the temperature to stabilize.
3. If a Viton A O-ring or other leak-free gasket is used
to connect the probe nozzle to the probe liner, leak check the
train at the sampling site by plugging the nozzle and pulling a
vacuum of 380 mm (15 in.) Hg. Note; A lower vacuum may be used
if it is not exceeded during the test.
4. If an asbestos string is used for the probe gasket, do
not connect the probe to the train; instead, first plug the inlet
-------
Section No. 3.9.4
Revision No. 0
Date January.4, 1982
Page 8 of 21
to the filter holder and pull a vacuum of 380 mm (15 in.) Hg (see
previous note), and then connect the probe to the train to leak
check at a vacuum of about 25 mm (1 in.) Hg.
Alternatively, the probe may be leak checked with the rest
of the sampling train in one step at a vacuum of 380 mm (15 in. )
Hg (APTD-05813 and APTD-05764).
1. Start the pump with the bypass valve fully open and the
coarse adjust valve closed.
2. Open the coarse adjust valve, and slowly close the
bypass valve until the desired vacuum is reached. Note: Do not
reverse the direction of the bypass valve; this will cause dis-
tilled water to back up from the impingers into the filter
holder. If the desired vacuum is exceeded, either leak check at
the higher vacuum or end the leak check (step 3 below) and start
over.
3. When the leak check is complete, slowly remove the plug
from the inlet to the probe or the filter holder; close the
coarse adjust valve; and immediately turn off the vacuum pump to
prevent the impinger water from being forced back into the filter
holder and to prevent the silica gel from being forced back into
the third impinger.
4. Visually check to be sure that water did not contact
the filter and that the filter has no tears before beginning the
test.
4.2.6.2 During the Sampling - If. a component (e.g., filter
assembly or impinger) change is necessary during the test, con-
duct a leak check before the change, according to the step-by-
step procedure outlined above.
1. Record the initial dry gas meter reading on Figure 4.2.
2. Be sure the vacuum is equal to or greater than the
maximum value recorded up to that point in the test.
3/ 3
3. If the leakage rate is £0.00057 m 7min (0.02 ft /min)
or 4% of the average sampling rate (whichever is less), the
results are acceptable, so no correction need be applied to the
total volume of dry gas metered.
o
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 9 of 21
4. If a higher leakage rate is obtained, either record the
leakage and correct the sample volume (Section 6.3(b) of the
Reference Method, Section 3.9.10), or void the sample run.
5. Record the final dry gas meter reading on Figure 4.2.
4.2.6.3 Posttest - (mandatory) At the conclusion of each sam-
pling run, conduct a leak check in accordance with the procedures
above.
1. Record the initial dry gas meter reading on Figure 4.2.
2. Be sure the vacuum is equal to or greater than the
maximum recorded during the sampling run.
3 3
3. If the leakage rate is <_0.00057 m /min (0.02 ft /min)
or 4% of the average sampling rate (whichever is less), the
results are acceptable, so no correction need be applied to the
total volume of dry gas metered.
4. If a higher leakage rate is obtained, either record the
leakage rate and correct the sample volume (Section 6.3(a) or
6.3(b) of the Reference Method, Section 3.9.10), or void the
sample run.
5. Record the dry gas meter reading on Figure 4.2.
4.2.7 Sampling Train Operation - Just before beginning the
sampling, clean the portholes to minimize the chance of sampling
any deposited materials. Verify that the probe and the filter
heating systems (if required) are up to the desired temperatures
and verify that the pitot tube and the nozzle are positioned
properly. Follow the procedure below for sampling:
1. Record the initial dry gas meter readings, barometric
pressure, and other data on Figure 4.2.
2. Position the tip of the probe at the first sampling
point with the nozzle tip pointing directly into the gas stream.
When in position, block off the open area around the probe and
the porthole to prevent flow disturbances and unrepresentative
dilution of the gas stream.
3. Turn on the pump, and immediately adjust the sample
flow to attain isokinetic conditions; maintain a sampling rate of
±10% of the isokinetic rate (unless otherwise specified by the
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 10 of 21
administrator), and adjust the rate at any sampling point if a
20% variation in velocity pressure occurs. Note: Do not exceed
the maximum AH. Use nomographs or programmed calculators to
rapidly determine the orifice pressure drop corresponding to the
isokinetic sampling rate. If the nomograph is designed as shown
in APTD-0576,4 use it only with a Type S pitot tube which has a
C coefficient of 0.85 ± 0.02 and only when the stack gas dry
molecular weight (M,) is 29 ± 4, if C_ and M_ are outside these
P
limits, do not use the nomograph without compensating for the
differences. Recalibrate the isokinetic rate or reset the nomo-
graph if the absolute stack temperature (T ) changes by >10%.
5
4. Take other required readings (Figure 4.2) at least once
at each sampling point during each time increment.
5. Record the final dry gas meter readings (Figure 4.2) at ^_^
the end of each time increment. ( )
6. Repeat steps 3 through 5 for each sampling point.
7. Turn off the pump; remove the probe from the stack;
record the final readings after each traverse.
8. Conduct the mandatory posttest leak check (Subsec-
tion 4.2.6.3) at the conclusion of the last traverse. Record any
leakage rate. Also, leak check the pitot lines (Method 2, Sec-
tion 3.1.2); the lines must pass this leak check to validate the
velocity pressure data.
9 Disconnect the probe, and then cap the nozzle and the
end of the probe with polyethylene or equivalent caps.
Periodically during the test, observe the connecting glass-
ware—from the probe, through the filter, to the first impinger—
for water condensation. If any is evident, adjust the probe
and/or filter heater setting upward until the condensation is
eliminated; add ice around the impingers to maintain the silica
gel exit temperature at 20°c (68°F).
The manometer level and zero should also be checked periodi-
cally during each traverse. Vibrations and temperature fluctua-
tions can cause the manometer zero to shift.
o
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 11 of 21
4.3 Sample Recovery
After the sampling is complete and the data are recorded for
all points, begin the cleanup procedure immediately.
1. Allow the probe to cool until it can be safely handled.
2. Wipe off all external particulate matter near the tip
of the probe nozzle.
3. Cap the tip loosely to keep from losing part of the
sample; capping it tightly while the sampling train is cooling
can cause a vacuum to form in the filter holder, and can cause
impinger water to.be drawn backward.
4. Remove the probe from the sample train before moving
the sample train to the cleanup site.
5. Wipe off the silicone grease, and cap the open outlet
of the probe; be careful not to lose any condensate that is
present.
6. Wipe off the silicone grease from the filter holder
inlet, and cap this inlet.
7. Remove the umbilical cord from the last impinger and
cap the impinger.
8. Wipe off the silicone grease and then cap off the
filter holder outlet and any open impinger inlets or outlets with
ground-glass stoppers, plastic caps, or serum caps.
9. Transfer the probe and the filter-impinger assembly to
an area that is clean and protected from the wind to minimize the
chances of contaminating or losing any of the sample. Inspect
the train before and during disassembly, and note any abnormal
conditions.
4.3.1 Probe, Filter, and Impinqer Catches - This step-by-step
procedure should be followed carefully to recover virtually all
of the sample collected in the probe, filter, and impinger.
1. Use a graduated cylinder to measure (to the nearest
1 ml) the volume of water in the first three impingers and any
condensate in the probe.
2. Record the values on the sample recovery and integrity
form Figure 4.3.
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 12 of 21
Plant Q-SxiTvv>L™jLL^vO SvwJI±Ln, Sample date T^O-u 7.
sj J
Sample location P
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 13 of 21
3. Transfer the impinger water from the graduated cylinder
into a polyethylene container.
4. Add the filter to this container using procedures
subject to the Administrator's approval.
5. Be sure that dust on the outside of the probe or other
component (e.g., the probe nozzle, probe fitting, probe liner,
first three impingers, impinger connectors, and filter holder)
does not get into the sample while cleaning other sample-exposed
surfaces with deionized distilled water; use <500 ml for the
entire wash, and add these, washings to the washings and the
filter in the polyethylene container. To this container add the
rinsings from the probe and nozzle, as described in the following
procedure.
Probe and Probe Nozzle - Having two people clean the probe
should minimize sample losses. Keep brushes clean and protected
from contamination at all times.
1. Carefully remove, the probe nozzle, use a Nylon bristle
brush to loosen particles from the inside surfaces; use a wash
bottle to rinse with deionized distilled water until no particles
are visible.
2. Brush and rinse the inside parts of the Swagelok fit-
ting with deionized distilled water in a similar way.
3. Rinse the probe liner by squirting deionized distilled
water into the upper end of the probe and by tilting and rotating
the probe so that all inside surfaces are wetted, and let the
water drain from the lower end through a funnel (glass or poly-
ethylene) and into the container.
4. Follow the rinse with a cleaning with a probe brush.
Hold the probe in an inclined position, and sguirt deionized
distilled water into the upper end while pushing the brush with a
twisting action through the probe and catching any water and
particulate matter that is brushed from the probe into the sample
container. Note: Brush three times, or at least six times for
stainless steel or other probes which have small crevices that
entrap particulate matter.
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 14 of 21
o
o
5. Rinse the brush with deionized distilled water, and
quantitatively collect these washings in the sample container.
6. After cleaning the brush, make a final rinse of the
probe by repeating steps 1-3.
Impinger and Filter Assembly - To recover fine particulates
from these components, brush, wash, and add the rinsings to the
container.
1. Rinse the inside surface of each of the first three
impingers and their connecting glassware three times using small
portions of deionized distilled water for each rinse, and brush
each sample-exposed surface with a Nylon bristle brush, to ensure
recovery of fine particulate matter. Make a final rinse of each
surface and the brush.
2. Be sure that all joints have been wiped clean of sili-
cone grease before brushing and rinsing with deionized distilled
water the inside of the filter holder (front-half only if after
the third impinger) three or more times as needed; make a final
rinse of the brush and filter holder. ;
Container - The following steps should be followed after all
water washings and particulate matter have been collected in the
sample container.
1. Tighten the lid so that water will not leak out when it
is shipped to the laboratory.
2. Mark the height of the fluid level so that the re-
ceivers can determine whether leakage has occurred during trans-
port.
3. Label the container clearly to identify its contents;
example sample label is shown in Figure 4.4.
4.3.2 Sample Blank - Prepare a blank by placing an unused filter
in a polyethylene container and by adding a volume of water equal
to the total volume in the average sample. Process the blank in
the same manner as the field samples. C j
4.3.3 Silica Gel - Note the color of the indicating silica gel
to determine whether it has been completely spent, and make a
notation of its condition on Figure 4.3. ' /\
• I -
v / -
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 15 of 21
Plant Alu/niwm 5/wt.rtA City C a past ^/fu, 77/vtf-
II //
Site S'we.rM Our<.er Sample type P/uoAiJt SAtnple
Date /--^-go Run number AS-J*
Front rinse G' Front filter S' Front solution D
Back rinse 0^ Back filter B' Back solution C3X
Solution Level marked
• •
Volume: Initial Final ^
— s-
Clean up by §
— ... , ^
Figure 4.4. Example of a sample label.
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 16 of 21
1. Transfer the silica gel from the fourth impinger to its
original container using a funnel and a rubber policeman, and
seal the container. It is not necessary to remove the small
amount of dust particles that may adhere to the impinger wall;
since the weight gain is used for moisture calculations, do not
use water or other liquids to transfer the silica gel.
2. Determine the final weight gain to the nearest 0.5 g>
if a balance is available.
4.4 Sample Logistics (Data) and Packing of Equipment
Follow the sample recovery procedures for the required
number of test runs, and record all data on Figure 4.3. If the
probe and the glassware (impinger, filter holder, and connectors)
are to be used in the next test, rinse all with distilled deion-
ized water and then acetone. To document the data and to prepare
the sample for shipping the following steps are recommended after
the test.
1. Check all sample containers for proper labeling (time,
date, and location of tests, number of tests, and any other
pertinent data).. Be sure a blank has been taken and labeled.
2. Duplicate all data recorded during the field test, to
avoid costly mistakes, by using either carbon paper or data forms
and a field laboratory notebook. Avoid using water soluble pens.
3. Mail one set of data to the base laboratory or give it
to another team member or to personnel in the agency; handcarry
the other set.
4. Examine all sample and blank containers and sampling
equipment for damage and for proper packing for shipment to the
base laboratory, and label all shipping containers to prevent
loss of samples or equipment.
5. Make quick checks of sampling and sample recovery
procedures by using the on-site checklist, Figure 4.5.
o
o
o
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 17 of 21
Apparatus
Probe nozzle: stainless steel >s glass
Button-hook \^ elbow size
Clean?
Probe liner: borosilicate quartz other
Clean?
Heating system*
Checked?
Pitot tube: Type S ix- . other
Properly attached to probe?*
Modifications
Pitot tube coefficient
Differential pressure gauge: two inclined manometers
other _ ^ sensitivity p.Ot - o to /
Filter holder: borosilicate glass ^ _ glass frit
filter support _ silicone gasket _ other
Clean? yS
_
Condenser : number of impingers
Clean?
Contents 1st /G£>/rjj #
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 18 of 21
-Figure 4.5 (continued)
Meter box leveled? />JL^ Periodically?
Manometers zeroed?
AH@ from most recent calibration /.
Nomograph setup properly?
Care taken to .avoid scraping nipple or stack wall?*
Effective seal around probe when in-stack?
Probe moved at proper time?
Nozzle and pitot tube parallel to stack wall at all times?*
Filter changed during run?
Any particulate lost? -T^LA
Data forms complete and data properly recorded?* _ _
Nomograph setting changed when stack temp changed significantly?
Velocity pressure and orifice pressure readings recorded
accurately?*
Sampling performed at a rate <1.0 cfm?
Posttest leak check performed?* LU^> u (mandatory)
Leakage rate o. o/ @ in. Hg /S" .w,..
Orsat analysis AXXI^ from stack integrated ^ ^^^
Fyrite combustion Analysis sample location I)
Bag system leakchecked?* u^u^> V /
If data forms cannot be copied, record:
approximate stack temp 31 7e>f volume metered 81
% isokinetic calculated at end of each run
SAMPLE RECOVERY
Brushes: nylon bristle (AJU^> other
Clean? /
Wash bottles;" polyethylene or glass
Clean?
Storage containers:polyethylene -^_;
Probe allowed to cool sufficiently? /^o-. <^X? 5"..
Cap placed over nozzle tip to prevent loss of particulate?*
During sampling train disassembly, are all openings capped?
Clean-up area description:
Clean? iM^* Protected from wind?
_ _
Filters: paper ° jue^^> ~ _ type
Silica gel: type" (6 to 16 mesh)? new? ^iJL^-J used?
Color? ^JjJLuut ^ _ Condition?Q Ci
Filter handling: tweezers used?
surgical gloves? _ other
Any fluoride spilled?*
(continued) )/
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 19 of 21
Figure 4.5 (continued)
Water distilled?
Stopcock grease: acetone-insoluble?
heat-stable silicone? ^
Probe handling: distilled water rinse
Fluoride recovery from: probe nozzle ^
probe fitting uu^ probe liner
(/other
front half of filter holder
Blank: filter
distilled water
Any visible particles on filter holder inside probe?:*
All jars adequately labeled? _
Liquid level marked on jars?*
Locked up?
Filter blank
Sealed tightly?
0
*Most significant items/parameters to be checked.
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 20 of 21
o
TABLE 4.1. ACTIVITY MATRIX FOR ON-SITE MEASUREMENT CHECKS
Apparatus
Sampling
Filter
Condenser
(addition of
reagents)
Assembling
sampling
train
Sampling
(isokineti-
cally)
Acceptance limits
Centered in holder; no
breaks, damage, or con-
tamination during
loading
100 ml of distilled
water in first two
impingers; 200-300 g
silica gel in fourth
impinger
of
1. Specifications
in Fig 1.1
2. Leak rate <4% of
sampling volume or
0.00057 mVmin (0.02
ftVmin), whichever is
less
1. Within ±10% of
isokinetic condition
and at a rate of less
than 1.0 ftVmin
2. Standard checked
for minimum sampling
time and volume; sam-
pling time >2 min/pt
Frequency and method
of measurements
Use tweezers or surg-
ical gloves to load
Use graduated cylinder
to add water, or weigh
each impinger and its
contents to the near-
est 0.5 g
1. Check specifica-
tions before each
sampling run
2. Leak check before
sampling by plugging
nozzle or inlet to
first impinger and
pulling a vacuum of
380 nnn (15 in.) Hg
1. Calculate for
each sample run
2. Make a quick cal-
culation before test,
and exact calculation
after
Action if
requirements
are not met
Discard fil-
ter, and
reload
Reassemble
system
O
1.
ble
2.
the
Reassem-
Correct
leak
1. Repeat
the test run
2. As above
(continued)
O
-------
Section No. 3.9.4
Revision No. 0
Date January 4, 1982
Page 21 of 21
TABLE 4.1 (continued)
Apparatus
Sample recovery
Sample
logistics,
data collec-
tion, and
packing of
equipment
Acceptance limits
3. Minimum number of
points specified by
Method 1
4. Leakage rate
<0.00057 mvmin (0.02
ftVmin) or 4% of the
average sampling vol-
ume, whichever is less
Noncontaminated sample
1. All data recorded
correctly
2. All equipment exam-
ined for damage and
labeled for shipment
3. All sample contain-
ers and blanks properly
labeled and packaged
Frequency and method
of measurements
3. Check before the
first test run by mea-
suring duct and using
Method 1
4. Leak check after
each test run or be-
fore equipment re-
placement during test
at the maximum vacuum
during the test (man-
datory)
Transfer sample to
labeled polyethylene
container after each
test run; mark level
of solution in the
container
1. After each test
and before packing
2. As above
3. Visually check
after each sampling
Action if
requirements
are not met
3. Repeat
the procedure
to comply
with specifi-
cations of
Method 1
4. Correct
the sample
volume or re-
peat the sam-
pling
Repeat the
sampling
1. Complete
the data
2. Repeat
the sampling
if damage
occurred ...dur-
ing the test
3. Correct
when possible
-------
o
o
o
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 1 of 19
5.0 POSTSAMPLING OPERATIONS
The postsampling operations include checks on the apparatus
used in the field during sampling to measure volumes, tempera-
tures, and pressures, and analyses of the samples collected in
the field and forwarded to the base laboratory. Table 5.1 at the
end of this section summarizes the quality assurance activities
for the postsampling operations.
5.1 Apparatus Checks
Posttest checks will have to be made on most of the sampling
apparatus. These checks will include three calibration runs at a
single orifice meter setting; cleaning; and/or routine mainte-
nance. . Cleaning and maintenance are discussed in Section 3.4.7
and in APTD-0576.4 Figure 5.1 should be used to record data from
the posttest checks.
5.1.1 Metering System - The metering system has two components
that must be checked-r-the dry .gas meter and the dry gas meter
thermometer(s).
The dry gas meter thermometer(s) should be compared with an
ASTM mercury-in-glass thermometer at room temperature. If the
two readings agree within ±6°C (10.8°F), the meter reading is
acceptable; if not, the meter thermometer must be recalibrated
(Subsection 2.2, Section 3.4.2) after the posttest check of the
dry gas meter. Use the higher meter thermometer reading (field
or recalibration value) in the calculations. If the field
readings are higher than the recalibration reading, no tempera-
ture correction is necessary; if the recalibration value is
higher, add the difference in the two readings to the average dry
gas meter temperature reading.
The posttest check of the dry gas meter is described in Sec-
tion 3.4.2. Any leaks in the metering system should have been
corrected before the posttest check. If the dry gas meter cali-
bration factor (Y) deviates by <_5% from the initial calibration
factor, the meter volumes obtained during the test series are
/•''" , *\f~> I
'/A^
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 2 of 19
o
Plant RlumiNUff) Smeller Calibrated by "7T /
Meter box number FB'I Date 4/- 3/...-.8O.
Dry Gas Meter
Pretest calibration factor, Y /D. Q86> • (within ±2%)
Posttest check, Y* . Q.38'7 (within ±5% of pretest)
Recalibration required? yes >S" no
If yes, recalibration factor, Y (within ±2%)
Lower calibration factor, Y &.4£6> for calculations (pretest or
posttest)
Dry Gas Meter Thermometers
Was a pretest temperature correction used? _^____ yes *s* no
If yes, temperature correction (within ±3°C (5.4°F) over
range)
Posttest comparison with mercury-in-glass thermometer?* (within
±6°C (10.8°F) at ambient temperature)
Recalibration required? | yes \s* no .—^
Recalibration temperature correction?"" (within ±3°C C ]
(5.4°F) over range)* v~x
If yes, no correction necessary for calculations if meter
thermometer temperature is higher; if calibration temperature
is higher, add correction to average meter temperature for
calculations
Stack Temperature Sensor
Was a pretest temperature correction used? yes */* • no
If yes, temperature correction °C (°F) (within ±1.5% of
readings in K (°R) over range)
Average stack temperature of compliance test, T 78O K
Temperature of reference thermometer or solution for recalibre
tion £-£$ K (@)* (within ±10% of T )
Temperature of stack thermometer for recalibration S"<28 K
Difference between reference and stack thermometer temperaturesT"'
AT ' o K (°R)
Do values agree within ±1.5%?* t^ yes no
If yes, no correction necessary for calculations
If no, calculations must be done twice—once with the recorded
values and once with the average stack temperature corrected to
correspond to the reference temperature differential (AT ),
both final result values must be reported since there is no way
to determine which is correct /^~\
Figure 5;1 Posttest calibration checks.
(continued) ^<
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 3 of 19
Figure 5.1 (continued)
Barometer
Was the pretest field barometer reading correct? tx"" yes _ no
Posttest comparison?* ^9. £-£" mm (in.) Hg (±2.5 mm (0.1 in.) Hg)
Was calibration required? _ _____ Yes ^ no
_ _____ _
If yes, no correction necessary for calculations when the field
barometer has a lower reading; if the mercury-in-glass reading
is lower, subtract the difference from the field data readings
for the calculation
*Most significant items/parameters to be checked.
-------
o
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 4 of 19
acceptable; if Y deviates by >5%, recalibrate the. metering system
(Section 3.9.2). In the calculations, use the calibration factor
(initial or recalibration) that yields the lower gas volume for
each test run.
5.1.2 Stack Temperature Sensors - The stack temperature sensors
(thermocouples and thermometers) should be compared with a refer-
ence thermometer or with a thermocouple if the temperature is
>405°C (761°F).
For thermocouple(s), compare the thermocouple and the ref-
erence thermometer readings at ambient temperature. If the
values agree within ±1.5% of the absolute temperature, the cali-
bration is valid; if not, recalibrate the thermocouple (Section
3.9.2) to determine the difference (AT ) in the absolute average
s
stack temperature 200°C (360°F) if T is between 200°C and 405°C.(360° and
5
751°F). Compare the stack thermometer with a thermocouple at a
temperature that is within ±10% of T if T^ is >405°C (761°F).
S S
If the absolute temperatures agree within ±1.5% the calibration
is valid; if not, determine the error AT^ to correct the average
stack temperature.
5.1.3 Barometer - The field barometer should be compared to the
mercury-in-glass barometer. If the readings agree within ±5 mm
(0.2 in.) Hg, the field readings are acceptable; if not, use the
lower value for the calculations. If the field readings are
lower than the mercury-in-glass readings, the field data are
acceptable; if not, use the difference in the two readings (the
adjusted barometric value) in the calculations.
5.2 Base Laboratory Analysis
All fluoride samples should be checked by the analyst upon
receipt in the base laboratory for identification and sample
o
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 5 of 19
integrity. Any losses should be noted on the analytical data form
(Figure 5.2). Either void the sample or correct the data using a
technique approved by the administrator. If a noticeable amount
of sample has been lost by leakage, the following procedure may
be used to correct the volume.
1. Mark the new liquid level on the sample container.
2. Treat the ' sample as described in Subsection 5.2.3 and
note the final dilution volume (V , ).
,-jf- so In
3. Add water up to the initial mark on the container,
transfer the water to a graduated cylinder and record the initial
sample volume (vsoini) in milliliters.
4. Add water to the new mark on the container. -Transfer
the water to a graduated cylinder, and record the final volume
(V , -) in millileters.
5. Correct the volume by using the following equation:
V = V Vsolni
soln1 soln V , f
solnf
where
V , , = sample volume to be used in the calculations, ml;
V . = total volume of solution in which fluoride is con-
soln
tained, ml;
V , . = initial volume added to the container in the field,
ml;
Vsolnf = f^nal v°lume removed from the container in the base
laboratory, ml.
6 . Both the corrected and uncorrected values should be
'Submitted in the test report to the agency.
This analytical method is based on measurement of the activ-
ity or concentrations of fluoride ions (F~) in aqueous samples by
use of an appropriate calibration curve. Fluoride activity
depends, however, upon the total ionic strength of the sample and
the electrode does not respond to fluorides which are bound or
complexed. This difficulty is largely overcome by adding a
buffer of high total ionic strength and by requiring preliminary
distillation to eliminate interferent ions. The sample response
-------
-1-er //
Date
U er
yes
Sample location
Samples identifiable
Ambient temperature £ Q. 5"° (L__
Temperature of calibration standards
Temperature of samples £.&. 5"° 0—
Li
Analyst
no All liquid levels at marks
Constant temperature bath used
Date calibration standards prepared
Sample
number
/9F-/
ftp-a
f)F-3
fiF-4
Sample
identification
number
AF - no
rtF-iAO
AF-J30
#F- 140
Total
volume of
sample,
(Vt), ml
/ooo
/OOO
1 f)f>£>
I ODO
Aliquot
total sam-
ple added
to still
(At), ml
/&&
J OO
/ 06
J 06
Diluted
volume of
distillate
collected
(Vd), ml
^?5^>
3.5'C>
3
380
£37
Concentration
of fluoride
from cali-
bration curve,
(M), molarity
f>. Oooo 7*t
& .660 1 2.
Q.OOG&'J'b
O. 0OOQ 19-
Total
weight of
fluoride
in sample
(Ft), mg
3.S-/S~
s~.<*qq
j. /.Per
^.^70
Total weight of fluoride in sample
Ft =
Signature of analyst
Signature of reviewer or supervisor
Remarks:
Figure 5.2 Fluoride analytical data sheet.
O
O
id rt < O
CD fT> H- ft
W H-
CS\ C_i H- O
PJ O 3
O 0 2
H,I± &:
p 2; o
M H O •
vi>^ •
CO
Ul
03
M
O
-------
Section No.. 3.9.5
Revision No. 0
Date January 4, 1982
Page 7 of 19
to the ion-specific electrode is also monitored by a standard
reference electrode and a modern pH meter that has an expanded
millivolt scale.
Procedures are -detailed herein for preparing reagents,
blanks, control samples, distillation aliguots, reference and
working standards (including serial dilutions), and an expanded
calibration curve and procedure for treating, separating, and
measuring the fluoride in samples.
5.2.1 Reagents - The following reagents are needed for the
analyses of fluoride samples.
1. Calcium oxide (CaO) - ACS reagent grade powder or ACS
certified grade containing £0.005% fluoride.
2. Phenolphthalein indicator - 0.1% in 1:1 ethanol-water
mixture (v/v).
3. Sodium hydroxide (NaOH) - Pellets, ACS reagent grade or
the equivalent.
4. Sulfuric acid (HgSO4) - Concentrated, ACS reagent grade
or the equivalent.
5. Filters - Whatman No. 541 or the equivalent.
6. Water - Deionized distilled to conform to ASTM specifi-
cation D1193-74, Type 3. The analyst may omit the Mn04 test for
oxidizable organic matter if high concentrations of organic
matter are not expected.
7. Total ionic strength adjustment buffer (TISAB) - Add
approximately 500 ml of distilled water to a l-£ beaker; to this
add 57 ml of concentrated glacial acetic acid, 58 g of sodium
chloride and 4 g of CDTA (cyclohexylenedinitrilotetraacetic
acid); and stir to dissolve. Place the beaker in a water bath
until it has cooled, and then slowly add about 150 ml of 5M NaOH,
while measuring the pH continuously with a calibrated pH elec-
trode and a reference electrode pair, until the pH is 5.3. Cool
to room temperature, pour into a l-£ volumetric flask and dilute
to the l-£ mark with distilled water.
8. Hydrochloric acid (HC1) - Concentrated ACS reagent
grade or the equivalent.
-------
O
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 8 of 19
9. Sodium fluoride (NaF) standard (0.1 M) - Dissolve 4.2 g
± 0.002 g ACS reagent grade NaF, which has been dried for a mini-
mum of 2 h at 110°C (230°F) and stored in a desiccator, in deion-
ized distilled water, and dilute to 1-2 with deionized distilled
water; this solution contains 0.1 M of Fluoride.
5.2.2 Blanks - The three blanks needed for the analysis are a
filter blank to ensure that the quality of the filter is accept-
able, a distillation blank to avoid cross contamination, and a
sample blank to analyze with the samples to verify the purity of
the reagents used in sampling and analyses.
-.,
1. Filter blanks - Determine the fluoride content of the
sampling filters upon receipt of each new lot and at least once
for each test series. Randomly select three filters from each
o
1. Add each filter to 500 ml of distilled water. ^-s
2. Treat the filters exactly like a sample (Subsec-
tion 5.2.3).
3. Use a 200 ml aliquot for distillation. Initially,
the filter blank must be <0.015 mg F/cm2; if not, reject this
batch and obtain a new supply of filters.
2. Distillation blank - Check the condition of the acid in
the distillation flask (Subsection 5.2.5) for cross-contamination
after every 10th sample by adding 220 ml of distilled water to
the still pot and then proceed with the analysis. If detectable
amounts of fluoride (>0.00001 M) are found in the blank, replace
the acid in the distillation flask.
3. Sample blank - Prepare the sample blanks in the field
at the same time and with the same reagents used for sample
recovery.
1. Add an unused filter from the same batch used in
sampling to a volume of distilled water equal to the average
amount used to recover the samples.
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 9 of 19
2. Treat the sample blank in the same manner as the
samples are treated (Subsection 5.2.3). Analyze the sample
blanks with the samples.
5.2.3 Sample Preparation - Use the following procedure to pre-
pare samples for distillation. Distillation is not required if
it can be shown to the satisfaction of the Administrator that
fluoride results are unaffected by the alternate analytical pro-
cedure (e.g., ash and fusion of particulate matter with subse-
quent ion selective electrode analysis, or direct electrode anal-
ysis of gases trapped in impingers).. - .
1. Filter the contents of the sample container (including
the sample filter) through a Whatman No. 541 filter or the equiv-
alent into a 1500-ml beaker; if the filtrate volume is >_900 ml,
add NaOH to make the filtrate basic to phenolphthalein, and then
evaporate to <900 ml.
2. Place the Whatman No. 541 filter containing the insolu-
bles (including the sample filter) in a nickle crucible, add a
few milliliters of water; and macerate the filter with a glass
rod.
3. Add 100 mg or sufficient quantity of CaO to the nickel
crucible to make the slurry basic; mix thoroughly; and add a cou-
ple drops of phenolphthalein indicator, which turns pink in a ba-
sic medium. Note: If the slurry does not remain basic (pink)
during/the evaporation of the water, fluoride will be lost; if
the slurry becomes colorless, it is acidic so add CaO until the
pink returns.
4. Place the crucible either in a hood area under infrared
lamps or on a hot plate at low heat (approximately 50-60°C)
(122-140°F), and evaporate the water completely; then place the
crucible on a hot plate under a hood and slowly increase the
temperature for several hours or until the filter is charred.
5. Place the crucible in a cold muffle furnace and gradu-
ally (to prevent smoking) increase the temperature to 600°C
(1112°F); maintain the temperature until the crucible contents
are reduced to an ash containing no organic material; and remove
the crucible from the furnace to cool. /'"""~~
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 10 of 19 ^.^
6. Add approximately 4 g of crushed NaOH pellets to the
crucible, and mix; return the crucible to the furnace, and fuse
the sample for 10 min at 600°C (1112°F); and then remove the
sample from the furnace, and cool it to ambient temperature.
7. Use several rinsings of warm distilled water to trans-
fer the contents of the crucible to the beaker containing the
filtrate (step 1) and finally, rinse the crucible with two 20-ml
portions of 25% (v/v) H2S04, and carefully add the rinses to the
beaker.
8. Mix well, and transfer the beaker contents to a 1-8,
volumetric flask. Record this volume as Vt on the data form
(Figure 5.2). Dilute to volume with distilled water, and mix
thoroughly; and allow any undissolved solids to settle.
9. Weigh the spent silica gel and report the weight to the
nearest 0.5 g on the sample integrity and recovery form.
5.2.4 Acid-water Ratio - The acid-water ratio in the distilla-
o
tion flask should be adjusted by following this procedure. Use a
protective shield when carrying out the procedure.
1. Place 400 ml of distilled water in the 1-2 distillation
flask, and add 200 ml of concentrated H2S04. Slowly add the
H2S04, while constantly swirling the flask.
2. Add soft glass beads and several small pieces of broken
glass tubing, and assemble the apparatus as shown in Figure 1.3.
3. Heat the flask until it reaches a temperature of 175°C
(347°F), and discard the distillate, and hold the flask for
fluoride separation by distillation.
5.2.5 Fluoride Separation (Distillation) - Fluoride in the
acid-water adjusted flask can be separated from other consti-
tuents in the aqueous sample by distilling fluosilicic (or hydro-
fluoric) acid from a solution of the sample in an acid with a
higher boiling point. Samples with low concentrations"of fluo-
ride (e.g., samples from an inlet and outlet of a^ scrubber) ^_^^
should be distilled first to eliminate contamination by carryover f J
of fluoride from the previous sample. If fluoride distillation
in the milligram range is to be followed by distillation in the
fractional milligram range, add 200 ml of deionized distilled /<_/
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 11 of 19
water and redistill similar to the acid adjustment procedure,
Subsection 5.2.4, to remove residual fluoride from the distilla-
tion system.
1. Cool the contents of the distillation flask (acid-water
adjusted) to <80°C (176°F).
2. Pipette an aliquot of sample containing <10.0 mg F into
the distilling flask, and add distilled water to make 220 ml.
The aliquot size (A, ) should be entered on the data form (Figure
5.2). Note; For an estimate of the aliquot size that contains
<_10 mg F, see Subsection 5.2.6.
3. Place a 250-ml volumetric flask at the condenser exit;
heat the distillation flask as rapidly as possible with a burner,
while moving the flame up and down the sides of the flask to
prevent bumping; conduct the distillation as rapidly as possible
(<_15 min). Slow distillations have been found to give low fluo-
ride recovery. Collect all distillate up to 175°C (347°F).
Caution; Heating >175°C (347°F) will cause H-SO^ to distill
over. Note; The H2SC- in the distilling flask can be reused
until carryover of interferent or until poor fluoride recovery is
shown in the distillation blanks and the control samples,
4. Before distilling samples and after every 10th sample,
distill a control sample to check the analytical procedures and
interferences (Subsection 5.2.6).
5.2.6 Control Sample - A control sample should be used to verify
the calibration curve and the analytical procedures before and
during the analysis of the field samples. Use the following
procedures.
1. The 0.05M NaF control sample stock solution - Add 2.10
g of reagent grade anhydrous NaF to a l-.fi. volumetric flask; add
enough distilled water to dissolve; and dilute to l-£ with dis-
tilled water.
2.. The 0.005M NaF working solution - Pipette 100 ml of the
0.05M NaF stock solution into a l-£ volumetric flask, and dilute
to the mark with distilled water to get the 0.005 M NaF working
solution. Note; The control should be within 0.004M and 0.006M
NaF; if not, take corrective action until these limits are met.
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 12 of 19
3. Analyze the working solution in the same manner as the
samples are analyzed (Subsections 5.2.5, 5.2.9, and 5.2.10).
5.2.7 Distillation Aliquot - The sample volume for distillation
should contain <10 mg F. Use the following procedure to estimate
the aliquot size.
1. Pipette a 25»ml aliquot of sample into a polyethylene
beaker.
2. Add an equal volume of TISAB buffer, and mix well.
3. Adjust the pH meter, and read the millivolts' for the
nondistilled sample and the calibration standard solutions (Sub-
section 5.2.8).
4. Determine the molarity of the nondistilled sample from
the calibration curve, and determine the size of the aliquot for
distillation by substituting the molarity (M) of the nondistilled
sample in the following equation: (J
.liquot for distillation (1, = e^LSdVmolarity (M)
The aliquot size is only an approximation since the interferring
ions have not been removed by distillation. If the estimate is
>220 ml, use 220 ml; if it is <220 ml, add distilled water to
make the total volume 220 ml; if required, dilute the sample to
get a minimum 1-ml aliquot.
5.2.8 Calibration Standards - Use the 0.1M NaF reference stan-
dard (Subsection 5.2.1) 'in the following procedure for preparing
serial dilutions.
1. Pipette 10 ml of 0.1M NaF into a 100-ml volumetric
flask, and dilute to volume with distilled water to get a 0.01M
standard.
2. Pipette 10 ml of the 0.01M standard solution to make a
0.001M solution in the same manner, and so on to make 0.0001M and
0.00001M solutions.
3. Pipette 50 ml of each of the standard solutions into ( J
separate polyethylene beakers, add 50 ml of TISAB buffer to each,
and mix well (50 ml of 0.01M diluted with 50 ml of TISAB is still
referred to as 0.01M). Prepare fresh 0.01M NaF standards daily.
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 13 of 19
A detailed explanation is in Section 3.9.2, Subsection 2.8 of
this method along with calibration curves (Figures 2.8 and 2.9 of
Section 3.9.2).
5.2.9 Sample Treatment - To treat the distilled fluoride in the
volumetric flask (from Subsection 5.2.5, step 3), follow this
procedure.
1. Dilute with distilled water to the 250-ml mark on the
volumetric flask at the condenser exit, and mix thoroughly.
2. Pipette 25-ml of the sample into a 50-ral volumetric
flask, dilute to the mark with TISAB buffer solution, and mix
well.
3. Bring the calibration standards and the samples to the
same temperature; if the ambient laboratory temperature fluctu-
ates more than ±2°C (4°F), condition the samples and standards in
a constant temperature bath.
5.2.10 Concentration Measurement - Some electrodes yield posi-
tive (direct F~ concentrations) and some yield negative (indi-
rect) values; if positive, recalibrate the electrode by using a
manufacturer-recommended standard, by adjusting the calibration
control (if needed) to the correct value, and by verifying the
calibration after measuring each standard and sample to prepare
the calibration curve.
Several precautions are needed before beginning the .proce-
dure.
1. Keep the pH meter on standby, and rinse between mea-
surements .
2. Keep the electrodes in the storage solution to prevent
overdrying if long periods of time are expected between uses.
3. Do not allow the electrode to touch the side of the
beaker during or between measurements.
4. Use of a stirrer will minimize electrode response time,
but stirring a solution before immersing the electrode may entrap
air around the crystal and cause needle fluctuations and errone-
ous readings.
-------
:' Section No. 3.9.5
Revision No.. 0
Date January 4, 1982
Page 14 of 19
Use an ion-specific electrode in the following procedure for
measuring the F~ concentration.
1. Transfer each standard and each sample to a series of
150-ml polyethylene beakers, and arrange each series so that the
lowest concentration will be read first to avoid carryovers.
2. Rotate the switch of the pH meter to standby, and allow
a 30-min warm-up period.
3. Raise the electrode from the storage solution in the
beaker, and rinse either the electrode thoroughly with distilled
water or soak the fluoride-sensing electrodes in distilled water
for 30 s before removing and blotting dry. Note: This step
should be done between each measurement.
4. Turn the adjustment knob to calibrate; immerse the
electrode in the NaF standard of lowest concentration.
5. Rotate the switch to millivolts (mV), and turn the ^^
adjustment knob to calibrate, read the millivolts of the known \ J
buffer solution from the meter, and record the value on Figure
5.2. Rotate the selector knob .to standby.
6. Raise the electrodes carefully from the buffer solu-
tion, and rinse thoroughly (step 3).
7. Immerse the electrodes carefully into a beaker of
standard solution, and set the beaker on a magnetic stirrer.
Note; If stirrer generates enough heat to change solution tem-
perature, place insulating material (e.g., cork) between the
stirrer and the beaker.
8. Rotate the selector knob to mV, read the mV from the
meter, .and record the value on Figure 5.2; allow the electrodes
to remain in the solution at least 3 min, and rotate the knob to
standby; take a final reading.
9. Repeat the above steps until all samples have been
read. Switch to standby, and then rinse and store the electrodes
in distilled water.
5.2.11 Expanded Calibration Curve - Use the following procedure
to construct an expanded calibration curve for analyzing samples
in the lower concentration range of <2 mg F/250 ml distillate and
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 15 of 19
for more accurate determinations of concentrations since samples
in the range are <0.001M NaF. Use this procedure to prepare
calibration standards, using the 0.1M NaF standard for serial
dilutions (Subsection 5.2.1).
1. Pipette 10 ml of the 0.1M NaF into a l-£ volumetric
flask, and dilute to volume using distilled water to get a 0.001M
standard.
2. Pipette 10 ml of the 0.001M standard, and dilute it
to 100 ml to make a 0.0001M standard.
3. Pipette 10 ml of the 0.0001M standard, and dilute to
100 ml to make a 0.00001M standard solution.
4. Pipette 50 ml of the 0.001M standard into a 100-ml
volumetric flask, and dilute to volume with distilled water to
get a 0.0005M standard.
5. Pipette 10 ml of the 0.0005M standard into a 100-ml
volumetric flask, and dilute to volume to make a 0.00005M stan-
dard.
6. Calibrate the electrode, and construct a calibration
curve (Subsection 5.2.8). Note; As shown in Figure 5.3, the
nominal concentrations of 0.00001M, 0.00005M, 0.0001M, 0.0005M,
and 0.001M NaF should be plotted on the log axis and the elec-
trode potentials (mV) are plotted on a linear scale.
Control samples are needed to verify the expanded calibra-
tion curve and the analytical procedure before and during the
analysis of the field samples. Use the 0.005M control sample
(Subsection 5.2.1) for the serial dilutions.
1. Pipette 5 ml of the 0.005M control sample into a 100-ml
volumetric flask, and dilute to volume to get a 0.00025M control.
2. Pipette 50 ml of the 0.00025M control into a poly-
ethylene beaker, add 50 ml of TISAB buffer, mix well, and use to
validate the calibration curve and to provide hourly checks on
the daily calibration.
3. Analyze the control sample (Subsection 5.2.5), and
record the data on the laboratory worksheet (Figure 5.4).
-------
i -
LU
I—
o
Q.
UJ
Q
O
o:
h-
o
LU
140
160
180
200
240
260
280
300
-/<"i V
-f*
Date ~if^_L L
Sample temp
Analyst
Reviewer
cQ.s-rc
0.00001
Molaritv
0.00001
0.00005
0.0001
0.0005
0.001
Control
sample
0.00005 0.0001
FLUORIDE HOLARITY, M
Figure 5.3. .Expanded fluoride calibration curve.
3 CO
al
0.0005
0.001
*TJ O JO trt
w p» n> n>
iQ fl" ^ •-•• o'
'H
U)
cn
CO
NJ
o
o
o
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 17 of 19
Date standards prepared
LABORATORY WORKSHEET
Date
/.
V
/.
Temperature of standards £0 .S~0 (2. Electrode number _____ OO I
Standard number
1
2
3
4
5
Control sample
Concentration, M
0.001
0.0005
0.0001
0.00005
0.00001
o.aooA^
Electrode potential, mV
£0
3A
A6>7
£30
3oo
ay-*-
Note: The control sample, from the calibration curve, must be between
0.0002M and 0.0003M.
Signature of analyst
Signature of reviewer
Figure 5.4. Expanded calibration curve data form.
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 18 of 19
o
Table 5.1 ACTIVITY MATRIX FOR POSTSAMPLING OPERATIONS
Apparatus
•Sampling
Apparatus
Dry gas meter
Meter thermome-
ters
Barometer
Stack tempera-
ture sensors
Base Laboratory
Analysis
Reagents
Acceptance limits
±5% of calibration
factor
±6°C (10.8°F) ambient
temperature
±5 mm (0.2 in.) at
ambient pressure
±1.5% of the reference
thermometer or thermo-
couple
Prepare according to
Subsec 5.2
Frequency and method
of measurements
Make three runs at a
single intermediate
orifice setting at
highest volume of
test (Sec 3.9.2)
Compare with ASTM
mercury-in-glass
thermometer after
each test
Compare with mercury-
in-glass barometer
after each test
Compare with ref-
erence temperature
after each run
Prepare a calibration
curve when preparing
new reagent
Action if
requirements
are not met
Recalibrate;
use factor
that gives
lower gas
volume
Recalibrate;
':se higher
temperature
for calcula-
tions
Recalibrate;
use lower
barometric
value for
calculations
Recalibrate;
calculate
with and
without tem-
perature cor-
rections
Prepare new
solutions and
calibration
curves
(continued)
O
o
-------
Section No. 3.9.5
Revision No. 0
Date January 4, 1982
Page 19 of 19
Table 5.1 (continued)
Apparatus
Acceptance limits
Frequency and method
of measurements
Action if
requirements
are not met
Control sample
±2% when run with
fluoride standards and
±10% when distilled
and run with field
samples
Prepare new controls
before and during
analysis of field
samples
Prepare new
solution and
calibration
curve, and/or
change dis-
tillate
solution
-------
o
o
o
-------
Section No. 3.9.6
Revision No. 0
Date January 4, 1982
Page 1 of 7
6.0 CALCULATIONS
Calculation errors due to procedural or mathematical mis-
takes can be a large part of total system error. Thus, each set
of calculations should be repeated or spotchecked, preferably by
a team member other than the one that performed the original
calculations. If a difference greater than a typical roundoff
error is detected, the calculations should be checked step-by-
step until the source of error is found and corrected.
A computer program is advantageous in reducing calculation
errors. If a standardized computer program is used, the original
data entry should be checked and if differences are observed, a
new computer run should be made.
Table 6.1 at the end of this section summarizes the quality
assurance activities for calculations. Retain at least one
significant digit beyond that of the acquired data. Roundoff
after the final calculations for each run or sample to two sig-
nificant digits, in accordance with ASTM 380-76. Record the
results on Figure 6.1A or 6.IB.
6.1 Nomenclature
Terms used in Equations 6-1 through 6-7 are defined here for
use in the Subsections that follow.
n
B
ws
tb
2 2
= Area of nozzle, cross-sectional, m (ft )
= Aliquot of total sample added to still, ml
= Water vapor in the gas stream, proportion by
volume
= Concentration of fluoride in stack gas corrected
to standard conditions of 20°C, 760 mm Hg (68°F,
29.92 in. Hg) on dry basis, mg/m (Ib/ft )
= Total weight of fluoride in sample, mg (Ib)
= Total weight of fluoride in sample blank, mg (Ib)
= Percent of isokinetic sampling, %
-------
Section No. 3.9.6
Revision No. 0
Date January 4, 1982 (j
Page 2 of 7
SAMPLE VOLUME (ENGLISH UNITS)
vm = £ 8. • &4 7 ft*, T = S 3 A - £ °R, PX3>. = <3 9 •
-------
Section No. 3-16.2
Date June 30, 1988
Page 12
Assemble the static dilution system (see Figure 4.3) and leak check the
system by plugging inlet to the dilution system, placing the discharge
line in a container of water, and pulling a vacuum of about 1 in. of Hg.
The system is leakless if no bubbles are released from the discharge line.
If the system is not leakless, find the leak and correct it.
Calculate as the amount of diluent gas needed to obtain the desired
dilution or calibration gas concentration. Meter the desired amount of
gas into the bag. If the purpose of the static dilution is to prepare a
lower concentration of calibration gas, the calibration gas should be
added to the bag using a gas tight syringe. Record the data on the static
dilution system data form, Figure 2.6 or similar form. If the purpose of
the static dilution system is to collect a diluted stack sample, the
calculated amount of diluent gas is added to the bag and the stack gas is
metered into the bag from the stack. To calibrate this system, the
calculated amount of diluent gas should be metered into the bag and then a
calibration gas should be metered into the bag with the flowmeter that is
to be used in the field. Record the data on the static dilution system
data form. Figure 2.6 or similar data form.
Analyze the diluted calibration gas and analyze a different calibration
gas that is in the range of the diluted calibration gas. The two gases
must agree within 10# of each other for the system to be acceptable.
o
o
2.3 Thermometer
The thermometers(s) on the metering systems and the sample probes and lines
should be initially compared with a mercury-in-glass thermometer that meets ASTM E-
1 No. 63C or 63F specifications:
1. Place the thermometer to be calibrated and the mercury-in-glass thermo-
meter in a bath of boiling water. Compare the readings after the bath
stabilizes and then record on the calibration data form. Figure 2.7 or
equivalent.
2. Allow both thermometers to come to room temperature. Compare the readings
after the thermometers stabilize.
3. The thermometer is acceptable if the values agree within 3°C (5.4°F) at
both points.
4. Prior to each field trip, compare the temperature reading of the mercury-
in-glass thermometer at room temperature with that of the thermometer that
is part of the metering system. If the values are not within 6°C (10.8°F)
of each other, replace or recalibrate the meter thermometer.
2.4 Barometer
The field barometer should be adjusted initially and before each test series
to agree within 2.54 mm (-0.1 in.) Hg with a mercury-in-glass barometer or with the
pressure value reported from a nearby National Weather Service Station and correct-
ed for elevation. The tester should be aware that the National Weather Service
readings are normally corrected to sea level; uncorrected readings should be x—x
obtained. The correction for the elevation difference between the weather station f j
and the sampling point should be applied at a rate of -2.5 mm Hg/30 m (-0.1 in. V_y
Hg/100 ft) elevation increase, or vice versa for elevation decrease.
-------
Section No. 3.16.2
Date June 30, 1988
Page 13
Date
Source flo^metJer number
Dry gas meter number QM~ l~7
Ambient temperature
Barometric press
Organic compound
Certified concen, (X) 2.2-/Q /ppmv
mm (in.) Hg
Calibrated by
Date source meter calibrated _
Date dry gas meter calibrated
Dry gas meter calib factor (Y)
Leak check for total system
Vacuum during leak check
Date of calibration curve
///#/ 88
~
/& /*.
Initial dry gas meter reading, L (ft3)
Final dry gas meter reading, L (ft3)
Volume of diluent gas metered, L (ft3)
Gas metered X calibration factor (Y),{V2}
Flowmeter sampling rate, L/min (cfm)
Sampling time, min
Sampling rate X sample time, L (ft3),{V1)
Dilution ratio
Injection time, 24h
Distance to peak, cm
Chart speed, cm/min
Retention time, min
Attenuation factor
Peak area or units
Peak area X attenuation factor
Measured concentration," ppmv
Calculated concentration,1* ppmv, {Cs}
Percent difference,0 %
RUN 1
RUN 2
RUN
IB3.
/3 Z-
J2-/.6>/3
O.
2.0
2-0
3.04-
&.//0
2-6.6
ID
30 5P
* See Figure 5-1 for calculation.
b Calculated concentration (Cs) =
X (Vt)
Va)
ppmv
Measured concent - Calculated concent
0 Percent difference,
X 100
Measured concentration
The percent difference must be less than 10 % absolute.
Figure 2.6. Static dilution data form.
31
-------
Date
to/to
fio/M
Reference
thermometer
type
*w
*%£
Calibr
thermc
type
tS&
~fat.fiw-
CeufilJL,
1
ated
meter
use
sh*
htefar
no.
WHZ
?-z
Ambien
refer"
<#F
MT-
t temper
calibr"
•fi-P
JO'F
Measurec
ature
differ0
'w
+-!'!=
values
Boi
refer*
wr
Z/2"/r
ling wat
calibrb
3&F
*ll'f
er
differc
i-i-fi
-\'F
Calibrator's
initials
ATA1
A77V
TJ o en
£8 P (D
Oq c* O
0) (D ct
Temperature reading of the reference thermometer in °C or °F.
Temperature reading of the thermometer being calibrated in °C or °F.
Difference between the reference thermometer and the calibrated thermometer. This difference must
be less than 3°C (5.4°F) for than initial calibration and 6°C (lO.lpF) for the calibration check.
Figure 2.7- Thermometer calibration fora.
O
o
CO «
o
- UJ
U3 ON
OO •
OOI\i
o
o
-------
Section No. 3.16.2
Date June 30, 1988
Page 15
Table 2.1. ACTIVITY MATRIX FOR CALIBRATION OF EQUIPMENT
Apparatus
Wet test meter
Dry gas meter,
mass flow meter,
and rotameters
Critical orifice
Dilution system
Thermometers
Barometer
Acceptance limits
Capacity of about 2
L/min and accuracy
within 12
Y£ = Y + 0.03Y at a
point greater than
the flow rate range
to be used
K'i = K ± 0.03K'
Measured value for
diluted and undi-
luted calibration
gas must agree
within 10*
Within 3°C (5.
of true value
Within 2.5 mm
{0.1 in.) Hg of
mercury-in-glass
barometer or weather
station value
Frequency and method
of measurement
Calibrate initially,
then yearly by
liquid displacement
Calibrate vs. wet,
dry, or bubble meter
upon receipt and
after each test
Calibrate vs. wet,
dry, or bubble meter
upon receipt and
after each test
Calibrate upon
receipt and prior to
each field test using
calibration gases
Calibrate initially
as a separate com-
ponent with mercury-
in-glass thermometer;
check before each
test against mercury-
in-glass thermometer
Calibrate initially
using mercury-in-
glass barometer;
check before and
after each test
Action if
requirements
are not met
Adjust until
specs are met, or
return to vendor
Repair and
then recalibrate,
or replace
Repair and
then recalibrate,
or replace
Correct problem
and rerun cali-
bration
Adjust or replace
Adjust to
agree with
certified
barometer
-------
o
o
o
-------
Section No. 3.16.3
Date June 30, 1988
Page 1
3.0 PRESAMPLING OPERATIONS
This section addresses two major areas of presampling operations for Method
18: (1) preparation for, performance of, and interpretation of results for the
preliminary survey and (2) preparation for the final sampling. This section de-
scribes the preliminary survey as it applies to Method 18; for additional general
information on preliminary surveys, see Section 3«0 of this Handbook. The quality
assurance activities for the preliminary survey activities and the presampling
activities for final testing are summarized in Tables 3«1 and 3«2, respectively, at
the end of this section.
3.1 Preliminary Survey Measurements
The preliminary survey measurements are needed to properly design the final
emission test sampling and analysis protocol. The primary objective of the prelim-
inary survey is to collect a preliminary survey sample for determining which sampl-
ing procedure is most appropriate and for developing the optimum analytical pro-
cedures. Using the preliminary survey sample, estimates of the source concentra-
tion are made and the major organic components in the gas stream are identified.
Also, any compounds that may interfere with the quantitation of the target anal-
yte(s) are identified and the appropriate changes in the analytical procedures are
made. Other measurements made during the preliminary survey include sampling site
dimensions and gas stream properties. The preliminary survey is also used to
obtain a description of the process being sampled, to determine sampling logistics,
and, when possible, to collect bulk process samples and use emission screening
techniques. Use the data form shown in Figure 3-1 to record the preliminary survey
information.
3.2 Preliminary Survey Preparation
This section addresses the equipment and preparatory activities needed to
conduct the preliminary survey. Figure 3.2 can serve as an equipment checklist,
packing list, and/or equipment status form for the preliminary survey.
3.2.1 Measurement of Flue Gas Properties - The apparatus that may be required to
supplement information obtained from plant personnel during the preliminary survey
concerning the moisture level, temperature, and static pressure of the source
should be prepared for the preliminary survey as follows:
Barometer - The field barometer should be compared with a mercury-in-glass
barometer or with a National Weather Service Station (see Subsection 2.4) reading
prior to each field test.
Net Bulb/Dry Bulb Thermometers - It is recommended that for sources with stack
temperatures at or below 59° C, wet bulb/dry bulb thermometers be used to determine
stack gas moisture content. The thermometers should be compared with a mercury-in-
glass thermometer at room temperature prior to each field trip. The wet bulb/dry
bulb measurement may also be used, with the prior approval of the Administrator, to
determine stack gas moisture for sources where the stack temperature exceeds 59° C.
Method 4 Equipment - For sources with stack temperatures above 59°C, Method 4
equipment is recommended to determine stack gas moisture content. Prepare the
. . /"/ ••? ('' )
'
-------
Section No. 3.16.3
Date June 30, 1988
Page 2
I. Name of company Date_
Address
Contacts Phone
Process to be sampled
Duct or vent to be sampled
II. Process description
Raw material
Products
Operating cycle
Check: Batch Continuous Cyclic_
Timing of batch or cycle
Best time to test
III. Sampling site
A. Description
Site description
o
Duct shape and size
Material
Wall thickness inches
Upstream distance inches _diameter
Downstream distance inches diameter
Size of port
Size of access area
Hazards Ambient temp_
B. Properties of gas stream
Temperature °C °F, Data source_
Velocity_ , Data source_
Static pressure inches H20, Data source_
Moisture content %, Data source
Particulate content , Data source
Gaseous components
N2 % Hydrocarbons (ppm) Toxics/Acids (ppm)
02 % H2S
CO % HC1 __
C02 % HF _
S02 % Other
Figure 3-1- Preliminary survey data sheet.
(Continued)
O
-------
Section No. 3.16.3
Date June 30, 1988
Page 3
Figure 3.1 (Continued)
Hydrocarbon components
ppm
ppm
ppm
ppm
ppm
ppm
C. Sampling considerations
Location to set up GC
Special hazards to be considered_
Power available at duct_
Power available for GC
Plant safety requirements
Vehicle traffic rules
Plant entry requirements
Security agreements_
Potential problems
Safety equipment (glasses, hard hats, shoes, etc.)
D. Site diagrams. (Attach additional sheets if required).
IV. On-site collection of preliminary survey samples
A. Evacuated flasks
Flasks have been cleaned, heated in furnace and purged
with nitrogen?
B.
(Continued)
Flask evacuated to the capacity of pump?
Filter end of probe placed at center of stack, probe
purged and sampled collected into flask until flask is at
stack pressure?
Stopcocks closed and taped?
Duct temperature and pressure recorded?
Purged flasks
Flasks cleaned and purged with nitrogen?
Filter end of probe placed into stack, sample purged for
2 to 5 min and then stopcocks closed?
Stopcocks taped to prevent leakage?
Duct temperature and pressure recorded?
Stability and adsorption checks conducted?^
-------
Section No. 3.16.3
Date June '30, 1988
Page 4
o
Figure 3^1 (Continued)
C. Flexible bags
Bags have been blanked checked and leak checked?_
Sampling system leak checked?
Filter end of probe placed into center of stack and sample
obtained at a proportional rate for appropriate amount of
time?
f>uct temperature, barometric pressure, ambient temperature,
flow rate, static pressure, and initial and final sampling
time recorded?
Analysis performed within 2 hr?
Stability and adsorption checks conducted?
D. Adsorption tubes
Proper adsorption tube(s) selected based on the likely
analytes?
Probe or adsorption tube placed into center of stack and
sample obtained at a constant rate with a calibrated
system for appropriate time based on the expected concen-
trations of analytes?
Total sample time and sample flow rate (or the number of
pump strokes), the barometric pressure, and ambient
temperature recorded?
O
Water vapor was less than 2% or measures were taken to
protect or increase the adsorption capacity of the
adsorption tube(s)?
E. Quality assurance performance audit samples
Quality assurance audit samples collected in the same
manner as the emission samples?
F. Bulk samples and screening techniques
Bulk emission sample(s) collected?
Bulk liquid sample(s) collected?
Detector tubes or other screening techniques used?
O
-------
Section No. 3.16.3
Date June 30, 1988
Page 5
Apparatus check
Moisture Determination
W Bulb/D Bulb
Checked
Barometer
Calibrated*
Method 4
Probe, heated &
leak checked
Imptngers
Meter system
calibrated*
Velocity Determination
Pitot Tube
Number
Length
Pressure Gauge
Manometer
Other
Stack Thermometer
Calibrated
Evacuated Flask
Evacuated Flasks
Number
Cleaned
Oven heated
N2 purged
Probes
Number
Cleaned
Glass wool
Suction bulb
Pump
Purged Flask
Flask
Number
Cleaned
Oven heated
N2 purged
Acceptable
Yes
No
Quantity
Required
Ready
Yes
No
Loaded and Packed
Yes
No
*Most significant items/parameters to be checked.
Figure 3-2. Preliminary survey preparations.
(Continued)
-------
Section No. 3.16.3
Date June 30, 1988
Page 6
o
Figure 3.2 (Continued)
Apparatus check
Purged Flask (continued)
Probe
Number
deemed
Glass wool
Vacuum Source
Pump
Squeeze bulb
Bag Sampling
Probe Liner
S steel
Glass
Teflon tube
Length
Meter System
Flowmeter*
Pump
Evacuated can
Charcoal tube
Sample line
Tedlar Bags
Number
Blank checked
Leak checked*
Heated Box
Number
Heat checked
Adsorption Tube
Probe
Heated
Checked
Nonheated
Glass
S steel
Filter
Sample Line
Type
Length
Checked*
Acceptable
Yes
No
Quantity
Required
Ready
Yes
No
Loaded and Packed
Yes
No
O
*Most significant items/parameters to be checked.
(Continued)
O
-------
Section No. 3.16.3
Date June 30, 1988
Page 7
Figure 3.2 (Continued)
Apparatus check
Adsorption Tube (continued)
Pump and Meters
Pump
Orifice
Calibrated*
Rotameter
Calibrated*
Timer
Adsorption Tubes
Tw>e
Bulk Samples
20-ml Jars .
Cleaned
Acceptable
Yes
No
Quantity
Required
Ready
Yes
No
Loaded and Packed
Yes
No
*Most significant items/parameters to be checked.
-------
Section No. 3.16.3
Date June 30, 1988
Page 8 /""X
equipment for sampling following the procedures described in Section 3'3-3 of this ^—
Handbook. Method 4 equipment may also be used to determine the stack gas moisture
for sources where the stack temperature is at or below 59°C.
S-Type Pttot Tube and Differential Pressure Gauge - Prepare the S-type pitot
tube and the differential pressure gauge for sampling following the procedures de-
scribed in Section 3'1«3 of this Handbook.
3.2.2 Glass Flask Sampling - The apparatus and reagents required for the collec-
tion of preliminary survey samples using glass sampling flasks are prepared as
described below. Alternative equipment found suitable may be used subject to the
approval of the Administrator.
Probe - If a heated probe is to be used for sampling, then the probe's heating
system should be checked to see that it is operating properly. The probe should be
cleaned internally by brushing first with tap water, then with deionized distilled
water, and finally with acetone. Allow the probe to air dry. The probe should be
sealed at the inlet end and checked for leaks by applying a vacuum of 380 mm (15
in.) Hg. See Subsection 1.0 for leak check procedure. The probe is considered
leak free under these conditions if no loss of vacuum is seen after one minute.
Any leaks detected should be corrected or the probe should be rejected. If the
probe has an external sheath, the integrity of the seal between the sheath and the
probe liner should be checked to ensure ambient air does not dilute the gas sample.
Teflon Tubing - Prepare sections of tubing for connections between the probe
and each flask (or bag or tube) that constitutes a preliminary survey sample col-
lection device. Clean the tubing using the procedure described above for the
probe.
Quick Connects - The quick connects should be new or cleaned according to the
manufacturer's recommendations. Leak check the quick connects as described in
Subsection 1.0.
Glass Sampling Flasks - Prepare the glass sampling flasks for collecting
preliminary survey samples as follows: Remove the stopcocks from both ends of the
flasks, and wipe the parts to remove any grease. Clean the stopcocks, barrels, and
receivers with chloroform. Clean all glass parts with a soap solution, then rinse
with tap water followed by deionized distilled water. Place the flasks in a cool
glass annealing furnace and heat the furnace to 550°C. Maintain the flasks in the
oven at this temperature for one hour. After one hour, shut off and open the
furnace to allow the flasks to cool. Return the Teflon stopcocks to the glass
flasks (if glass stopcocks are used, apply a light coating of vacuum grease to the
stopcocks before returning to the flasks.) With both stopcocks open, purge each
assembled flask with high purity nitrogen for 2 to 5 minutes. Close off the outlet
stopcock followed by the inlet stopcock to maintain a slight positive nitrogen
pressure in the flask. Secure the stopcocks with tape to prevent them from opening
accidentally.
High-Vacuum Pump - A high-vacuum pump will be required for preliminary survey
sample collection using the evacuated flask procedure. Check the operation of the
pump prior to going to the field as follows: Check for minimum pump vacuum of 75
mm (3 in.) Hg absolute by attaching a Hg-filled U-tube manometer to the pump inlet
o
o
-------
Section No. 3.16.3
Date June 30, 1988
Page 9
and turning on the pump. If the minimum vacuum cannot be reached, then repair or
replace the pump.
Rubber Suction Bulb - A rubber suction bulb will be required for preliminary
survey sample collection using the purged flask procedure. The rubber suction bulb
should be checked for proper operation prior to going to the field as follows:
Attach the bulb to a water manometer, or equivalent, and squeeze the bulb until a
vacuum of at least 250 mm (10 in.) H20 is reached. Repair or replace the squeeze
bulb if sufficient vacuum cannot be developed.
3.2.3 Flexible. Bag Procedure - A flexible bag of Tedlar or aluminized Mylar can be
used to collect preliminary survey samples. If it is anticipated that Tedlar bags
will be selected as the final sampling method, then it is recommended that the
preliminary survey sample be collected using a Tedlar bag. In addition to the
apparatus described in Subsection 3.2.2 for the glass flasks (with the exception of
the flasks and the high-vacuum pump) the apparatus listed below will be required
and should be prepared as follows:
Tedlar or Aluminized Mylar Bags - Prepare new bags for preliminary survey
sampling by leak checking the bags before going to the field. The bags should also
be checked for contamination by filling with hydrocarbon-free air or nitrogen
during the leak check. The bags are checked as follows: Connect a water manome-
ter, or equivalent, using a tee connector, between the check valve quick connect on
the bag and a pressure source (or hydrocarbon-free air or nitrogen for conducting
the contamination check). Pressurize the bag to 5 to 10 cm (2 to 4 in.) H20. Loss
of pressure over a 30-second period indicates a leak. Alternatively, leave the bag
pressurized overnight: a deflated bag the following day is indicative of a leak.
Reject or repair any bags with leaks. After the hydrocarbon-free air or nitrogen
has remained in the bag for 24 hours, analyze the bag contents using a GC with a
flame ionization detector on the most sensitive setting. The bag should be re-
jected if any organic compounds are detected. If any organic compounds are
detected, the bags may be used if they are not the compounds to be sampled and
analyzed.
Rigid Leak-Proof Containers - Rigid containers can be used with the bags for
collecting preliminary survey samples. The rigid containers used to contain the
Tedlar bags during sampling should be checked• for leaks prior to going to the
field. The container should be leak checked with the bag in place as follows:
Using a tee connector, connect a water manometer or equivalent, between a pressure
source and the container outlet. Pressurize the container to 5 to 10 cm (2 to 4
in.) Hg. Any loss of pressure after 30 seconds indicates a leak. Reject or repair
the rigid container if a leak is indicated.
Direct Pump Sampling System - A direct pump sampling system can be used in
place of the rigid containers for collecting preliminary survey samples. If this
method is selected, then the system should be assembled and leak checked prior to
going to the field as follows: Assemble the system (see Figure 4.5). Bypass the
Tedlar bag and its protective container by attaching the vacuum line directly after
the rotameter using the quick connects on the sample and vacuum lines. Plug the
probe inlet and turn on the vacuum pump. If the system is leak free, the
rotameter should eventually indicate no flow. Alternatively, the sample line that
is attached to the sample bag can be placed in water. If bubbling stops, then the
system is leak free.
-------
Section No. 3-16.3
Date June 30. 1988
Page 10
Needle Valve and Eotameter - Prior to each field trip or at any sign of er-
ratic behavior, the flow control valve and the rotameter should be cleaned accor-
ding to the maintenance procedure recommended by the manufacturer.
3.2.4 Adsorption Tube Sampling - The adsorption tube sampling procedure can also
be used to collect the preliminary survey sample. If it is anticipated that ad-
sorption tubes will be selected as the final sampling method, then it is recom-
mended that the preliminary survey samples be collected using tubes containing each
potential type of adsorbent. In addition to the apparatus described in Section
3.2.1 for the glass flasks (with the exception of the flasks and a high-vacuum
pump) the apparatus listed below will be required and should be prepared as fol-
lows:
Adsorption Tubes - Check to see that the proper type of tube has been obtained
for collecting the target organic compounds. Refer to Table B in the Methods High-
lights Section to determine the proper adsorption material. Check to see that the
supply of adsorption tubes is sufficient to conduct the emission test, including
field blanks and desorption efficiency determinations.
Personnel Sampling Pump - A personnel sampling pump is used to collect the
adsorption tube samples. The pump should be calibrated following the procedures
described in Subsection 2.1.4.
Extraction Solvents - An extraction solvent will be required to prepare the
preliminary survey adsorption tube sample(s) for analysis. Refer to Table B in the
Methods Highlights Section to determine the proper extraction solvent.
3.3 Preliminary Survey Sample Collection
The preliminary survey sample collection includes flue gas or duct moisture
and velocity determinations in addition to collection of actual flue gas or duct
samples.
3.3-1 Preliminary Survey Moisture Determination - If the moisture content of the
flue gas in the duct to be tested cannot be obtained from the plant personnel, it
is determined using either wet bulb/dry bulb thermometers or Method 4 sampling
apparatus, depending on the flue gas temperature. If the flue gas temperature
cannot be obtained from plant personnel, then determine the flue gas temperature
using a calibrated thermocouple, thermometer, or equivalent temperature measuring
device.
Net Bulb/Dry Bulb Procedure - For flue gas streams at or below 59° C, the
moisture content of the flue gas should be determined using wet bulb/dry bulb
thermometers and the partial pressure equation shown below. Obtain the wet
bulb/dry bulb temperatures as follows:
1. Moisten the wet bulb thermometer wick with deionized distilled water.
2. Insert the thermometers into the flue gas stream and monitor the wet bulb
temperature.
3. When the wet bulb temperature has stabilized, record both the wet bulb and
dry bulb thermometer temperatures.
4. Calculate the flue gas moisture content using the equations below.
o
o
-------
(t
-(31AA/(Tw*390.86)
Section No. 3.16.3
Date June 30, 1988
Page 11
Equation 3-1
w2 =
%U20 = w2 - (0.00036? x (Td-Tw) x (l+(Tw-32)/1571)) x 100
where
Equation 3-2
w =
P. =
Calculated constant, saturation % H20 at Tw
Wet bulb temperature, °F,
Dry bulb temperature, °F,
Barometric pressure, in. Hg, and
Static pressure of duct, in.
H20.
Method 4 Moisture Procedure - Follow the procedure for Method
Section 3-3 of this Handbook.
described in
Method 2 Velocity Procedure - Follow the procedure for Method 2 described in
Section 3-1 of this Handbook to determine the flue gas or duct velocity at the
sampling point. If the velocity varies by more than 10% during the projected
sample run time, then proportional sampling will be required as described in Sub-
section 4.0. Because of the small size of some ducts, Methods 2A, 2C, or 2D may
have to be used. Follow the criteria and procedures described in the applicable
method.
3.3.2 Collection of Samples with Glass Sampling Flasks - Using the precleaned
glass sampling flasks, preliminary survey samples are collected using the evacuated
flask procedure or the purged flask procedure.
Evacuated Flask Procedure - Collect preliminary survey samples using the
evacuated flask procedure as follows:
1. Using a high-vacuum pump which is connected to one stopcock while the
other stopcock remains closed, evacuate each precleaned flask to the
capacity of the pump. A mercury manometer can be connected between the
pump and the flask using a tee connector to indicate when the maximum
vacuum is achieved. At this point, record the vacuum, and close off the
stopcock leading to the pump.
2. Remove the tubing leading to the pump and attach a glass tee (6-mm out-
side diameter, or equivalent) to the flask inlet with a short piece of
Teflon tubing.
3. Connect the end of the sampling probe to the glass tee using a short
length of Teflon tubing. The tubing must be of sufficient length to
reach the sampling point at the centroid of or no closer than 1 meter to
the duct wall.
4. Connect the rubber suction bulb to the third leg of the tee with a piece
of Teflon tubing or suitable flexible tubing.
5. Place a plug of glass wool in the probe inlet, enlarged to approximately
12-mm outside diameter, to serve as a filter to remove particulate mat-
ter.
6. Place the inlet (filtered) end of the probe at the sampling point and
purge the probe and sample line by repeatedly squeezing the rubber suc-
tion bulb until at least 7 air changes of the probe and sample line have
occurred.
-------
Section No. 3.16.3
Date June 30, 1988
: Page 12
7. After the probe and the sample line are completely purged, leave the
squeeze bulb in place, and open the inlet stopcock of the sampling flask
8. Leave the inlet stopcock open until the pressure in the sampling flask.
reaches the duct pressure. This should take about 15 seconds. Close the
inlet stopcock.
9. Remove the probe from the duct and disconnect the glass tee from the
flask.
10. Taps the stopcocks closed and label the flask with the plant name, date,
and sampling location, time, and sampling personnel.
11. Immediately after sampling, determine the flue gas temperature with a
calibrated thermocouple, thermometer, or equivalent temperature measuring
device, and determine the static pressure of the duct and the velocity
over a period of time equal to the predicted sample run time following
the procedures described in Section 3-1 *n this Handbook.
Purged Flask Procedure - Collect preliminary survey samples using the purged
flask procedure as follows: ,
1. Connect the small end of the sampling probe, of sufficient length to
reach the centroid of the duct to be sampled, to the inlet stopcock of a
precleaned glass sampling flask a sufficient length of Teflon tubing.
2. Connect the rubber suction bulb to the other stopcock with a piece of
Teflon tubing or suitable flexible tubing.
3. Place a plug of glass wool in the probe inlet, enlarged to approximately
12-mm OD, to serve as a filter to remove particulate matter.
4. Place the inlet (filtered) end of the probe at the centroid of or no
closer than 1 meter to the duct wall.
5- Purge the probe, sample line, and sample flask by repeatedly squeezing
the rubber suction bulb until approximately 7 air changes of the system
have occurred.
6. After the probe, sample line, and flask are completely purged, close off
the stopcock near the suction bulb, and then close off the stopcock con-
nected to the probe.
7. Remove the probe from the duct, and disconnect both the probe and the
suction bulb from the flask.
8. Tape the stopcocks closed and label the flask with the plant name, date,
and sampling location, time, and sampling personnel.
9. Immediately after sampling, determine the flue gas temperature with a
calibrated thermocouple, thermometer, or equivalent temperature measuring
device, and determine the static pressure of the duct and the velocity
over a period of time equal to the predicted sample run time following
the procedures described in Section 3*1 in this Handbook.
3.3-3 Flexible Bag Procedure - The flexible bags used to collect preliminary
survey samples must be leak checked and demonstrated to be free of contamination
following the procedure described in Subsection 3-2.2. The preliminary survey
sample collection using flexible bags can be conducted at a constant rate following
the procedure described in Subsection 4.3 for the evacuated container sampling
procedure, the direct pump sampling procedure, or, in explosive areas, the explo-
sion risk area sampling procedure. The flue gas or duct velocity and other
process parameters should be determined for designing the final sampling proced-
ures.
o
o
o
-------
Section No. 3.16.3
Date June 30, 1988
Page 13
3-3-4 Adsorption Tube Procedure - The adsorption tubes used to collect the pre-
liminary survey sample(s) should contain the adsorbent specified in Table B of the
Method Highlights Section for the target analyte(s). The sampling time or total
sample volume for the adsorption tube should be calculated based on expected con-
centration (s) of the volatile organic(s) present and the recommended capacity of
the adsorption media. Refer to the appropriate reference given in Table B to
determine the recommended sample volume taking into consideration the amount of
adsorbent to be used. For compounds not referenced in Table B, use a reference for
a compound with similar chemical characteristics. If the target analytes require
different adsorption media, then it is recommended that preliminary survey samples
be collected using each type of adsorbent. In the case where the compound is
unlike any other documented compounds, use two adsorption tubes connected in ser-
ies. Once a recommended volume is established, it is recommended that two
additional samples be collected with sample volumes one half and twice the
recommended volume. The procedure for collecting preliminary survey adsorption
tube samples is as follows:
1. Open the adsorption tube, and connect the primary tube section (large
section of adsorbent) to the sampling probe using a minimum length of
Teflon tubing or other nonreactive tubing.
2. Connect the outlet (backup section) of the tube to the next tube in
series, if additional adsorption capacity is required.
3. Connect the outlet of the last tube to the inlet of the calibrated per-
sonnel sampling pump using a sufficient length of tubing.
4. Insert the probe into the stack or duct and turn on the pump. Maintain
the adsorption tubes in a vertical position during sampling to prevent
channeling. Sample the gas stream for the time required to obtain the
optimal volume determined from the referenced method.
5. Immediately after sampling is completed, disconnect the tubes from the
tubing and seal the tube ends with teflon tape and plastic caps. Label
the tubes and store each tube in a screw cap culture tube or similar
container to protect them during shipment.
6. Record the total sampling time, the sample flow rate, the barometric
pressure, and the ambient temperature.
3-4 Preliminary Survey Sample Analysis and Interpretation
With the exception of the analysis of the glass sampling flasks, the analysis
of preliminary survey samples should follow the procedures described in Subsection
5-0. The analysis of the glass sampling flasks are described below (see Subsection
3-4.2). The analysis of preliminary survey samples is used to optimize the
analytical procedures and select the most appropriate sampling technique for final
sampling. Using Table C the Method Highlights Section, choose appropriate GC
detector(s). Based on the sampling technique(s) used to collect the preliminary
sample, choose a GC column from the selections listed in Table D of the Method
Highlights Section; the technical service department of column manufacturers or
plant laboratory personnel may also be consulted for additional suggestions on
column type(s). For glass flask samples and Tedlar or Mylar bag samples, use
calibration gas cylinders or calibration standards prepared in Tedlar bags. For
adsorption tube samples, prepare the calibration standards directly in the
desorption liquid(s) or on adsorption tube material(s) used to collect the samples.
3.4.1 Calibration Standards for Preliminary Survey Samples - Prepare a minimum of
three calibration standards for each compound of interest. The standards should
-------
Section No. 3.16.3
Date June 30, 1988
Page 14
cover a linear range for the particular GO detector, with the lowest standard and
the highest standard bracketing and a midrange standard approximating the expected
sample concentration. To estimate the sample concentration, prepare a preliminary
survey sample and perform a single analysis of the sample following the procedure
in the appropriate subsections below. During this sample analysis, determine if
adequate resolution has been achieved for each peak with a peak area greater than
5# of the total chromatographic peak area (excluding the desorption solvent peak)
using one of the procedures described in Subsection 3-4.1. Adequate resolution of
sample peaks will only be necessary in the chromatographic region(s) where the
target compound(s) are expected to elute. The GC analysis conditions and/or column
can be changed to achieve adequate resolution. The use of two different columns
may be necessary to ensure accurate identification of the gases. For analysis of
more than one target compound in very complex sample matrices, more than one
analysis using different GC conditions and/or columns may be required to achieve
adequate resolution for all target compounds.
For analysis of flask samples or bag samples, (!) use cylinder gases directly
(if available) or by dilution following the procedures described in Subsection
5-1.1 and 5-1.2, respectively, or (2) prepare standards in Tedlar bags following
the procedure described in Subsection 5-1-3 for gaseous materials or the procedure
described in Subsection 5-1-4 for liquid materials. For analysis of adsorption
tube samples, prepare calibration standards following the procedure described in
Subsection 5-1-6. Data forms should be used for recording calibration standard
preparation and analysis data (see Figures 5-4, 5-6, 5-8, and 5-9) and preliminary
survey sample analysis data (see Figure 5-1)-
The gaseous calibration standards for bag samples must be injected into the GC
using a gas sampling valve equipped with a stainless steel or Teflon sample loop
following the procedures described in Subsection 5-1 appropriate for the particular
type of gaseous standard used. Liquid calibration standards for adsorption tube
analysis must be injected into a heated sample injection port following the proced-
ure described in Subsection 5-1-6. The gaseous standards for glass flask samples
can be injected into the GC using either a gas sample valve, following the
appropriate procedure in Subsection 5-1 for the particular gaseous standard used,
or a heated injection port using a gas tight syringe following the procedure
described below; the same injection procedure used for the standards must be used
for the flask samples.
The procedure for injecting gaseous calibration standards using a gas tight
syringe is as follows:
1. Attach a GC septum to a piece of Teflon tubing and attach the tubing to
the outlet of the calibration gas cylinder regulators or the Tedlar bags
containing the calibration gases.
2. Insert the needle of the syringe through the septum, and repeatedly purge
the syringe by repeatedly filling and emptying the syringe 7 times.
3. After purging the syringe, fill the syringe past the mark corresponding
to the desired amount to be injected, and withdraw the syringe from the
septum. Stick the needle into a rubber stopper or a thick septum to
prevent dilution of the standard by ambient air.
4. Immediately before injecting the standard, remove the needle from the
stopper or septum, adjust the syringe to the desired volume, and inject
the standard into the heated injection port on the GC. Note the time of
injection on the strip chart and/or actuate the electronic integrator.
5. Repeat the injection of the standard until the peak areas from consecu-
tive injections agree within 5# of their average value.
o
o
-------
Section No. 3.16.3
Date June 30, 1988
Page 15
Perform initial tests using the calibration standards to determine the optimum
GC conditions to minimize analysis time while still maintaining sufficient resolu-
tion. Sufficient resolution can be determined following the procedure described by
Knoll42 or in EPA Method 625A/* where the baseline to valley height between two
adjacent peaks must be less than 25% of the sum of the two peak heights (see Figure
3-3).
Analyze the calibration standards, starting with the lowest concentrations
first. Repeat each standard analysis until two consecutive injections give indi-
vidual area counts within 5# of their average. Multiply the average of the re-
sponse for the two acceptable consecutive injections of each standard by the detec-
tor attenuation to determine the calibration area value for each standard. Record
the retention time for each compound and the calibration area for each standard.
Record the detector settings, the recorder/integrator attenuation for each
standard, the chart speed, the GC temperature settings, the column parameters (type
and length), and the carrier gas flow rate. Plot the concentration of the stan-
dards on the abscissa (x-axis) and the calibration area for each standard on the
ordinate. Perform a regression analysis, and draw the least squares line on the
plot. It is recommended, but not required for preliminary survey sample analysis,
that the validity of the calibration curve be checked using the audit procedures
described in Section 8.0. The audit sample may be analyzed at this time in lieu of
analysis during the final sample analysis with the prior approval of the
Administrator.
If positive identification of a target compound cannot be made by comparison
of the compound retention time to the retention time of one of the standards, then
use of a different type of column may be helpful. If positive identification still
cannot be achieved, then GC/mass spectrometry (GC/MS) or GC/infrared (GC/IR) tech-
niques should be used, with GC/MS recommended. In addition, any compounds, not
identified as target compounds, with peak areas greater than 5# of the total chro-
matographable peak area (excluding the solvent peak area for adsorption tubes)
should be identified by comparison to known standards or by using GC/MS.
3.4.2 Glass Flask Preliminary Sample Analysis - Since glass sampling flasks are
only used for preliminary survey samples, the analysis of the flasks is described
in this section. Glass sampling flasks require some pressurization prior to analy-
sis to withdraw the sample.
Using the ideal gas law, the amount of dilution of the sample that results
from pressurization can be estimated with enough accuracy to permit interpretation
of the preliminary survey sample results. The procedure for pressurizing a flask
is as follows:
1. Note if any condensation has collected in the flask. If it has, heat the
flask to the flue gas or duct temperature with an oven, heating tape, or
a heat lamp. Note; The pressurization of sealed glass containers by
heating is an inherently hazardous process. The use of a protective
shield to protect personnel from flying glass in the event of an
explosion is highly recommended. In addition, the flask should be
wrapped in cloth or other cushioning media during these operations.
2. Connect one end of the flask to a mercury manometer, open the stopcock,
and determine the initial pressure of the flask (Pi). Record Pi and the
initial absolute flask temperature (Tj) in °R or °K.
3. Connect the other end of the flask to a source of hydrocarbon-free nitro-
gen or air, and open the stopcock. Slowly pressurize the flask to a
maximum of 15 psig, and close the stopcock. Determine the final pressure
of the flask (Pf) and the final absolute temperature of the flask (Tx).
-------
1
w
v
Figure 3-3- Diagram showing EPA Method 625 criterion for adequate resolution of
overlapping compounds with similar mass spectra.
O
O
CD p CD
Oq rt O
CD CD ct
H-
M C_| O
cr>c 3
CD Z
O
UJ •
O
VO (T>
03-
OOUO
O
-------
Section No. 3-16.3
Date June 30, 1988
Page 1?
Note; The pressurization of sealed glass containers is an inherently
hazardous process. The use of a protective shield to protect personnel
from flying glass in the event of an explosion is highly recommended. In
addition, the flask should be wrapped in cloth or other cushioning media
during these operations.
Calculate the first dilution factor (Dj) using the following formula:
Di =
P x T
rf x ij
T x
i x
Equation 3-2
5- Allow the flask to equilibrate for 10 minutes. Note if any condensation
has formed. If condensation has formed in the flask and the flask did not
initially required heating, heat the flask to a temperature sufficient to
vaporize the condensate. If the condensate cannot be vaporized or if the
flask was already heated and more condensate formed during pressuriza-
tion, the sample cannot be analyzed accurately.
6. Close the stopcocks and disconnect the manometer and dilution gas.
Analyze the contents of a pressurized flask using a sample introduced into the
GC via a gas sampling valve by the following procedure:
1. Connect the sample flask to the injection valve with the valve in the
load position.
2. Open the stopcock connected to the valve, and allow the gas sample to
flow through the sample loop at 100 ml/min for 30 seconds (determined
with a rotameter connected to the outlet of the sample loop) or purge
with 5 times the sample loop volume, whichever is less. Close the stop-
cock, and allow the sample loop to return to ambient pressure.
3. Actuate the sample valve to inject the sample and record the injection
time.
4. Examine the chromatogram and determine if adequate resolution has been
achieved between individual target compound peaks and between target
compound peaks and any interfering compound peak with an area greater
than 5# of the total area of all peaks (excluding the desorption solvent
peak) using the procedure described in Subsection 3-4.1.
5. Determine the retention time for each peak by dividing the distance of
the peak maximum from the injection point by the chart speed.
6. Repeat the analysis, and determine the peak area and retention time for
each target compound identified during the second analysis. Although not
required for the preliminary survey sample analysis, the peak areas for
each target compound from consecutive injections should agree within 5%
of the average peak area. The retention times between the two injections
should agree within 0.5 seconds or 1% of the adjusted retention time
(compound retention time minus the time of elution of unretained peaks),
whichever is greater.
Analyze the contents of a pressurized flask using a sample introduced into the
GC via a gas tight syringe and a heated injection port sample by the following
procedure:
1. Attach a GC septum to one of the stopcocks on the glass flask. (Note:
Glass sampling flasks can be purchased with an integral septum porti)
2 Insert the needle of the syringe through the septum, and purge the sy-
ringe by repeatedly filling and emptying the syringe 7 times.
3. After purging the syringe, fill the syringe past the mark corresponding
to the desired amount to be injected, and withdraw the syringe from the
-------
o
Section No. 3.16.3
Date June 30, 1988
Page 18
septum. Stick the needle into a rubber stopper or a thick septum to
prevent dilution of the standard by ambient air.
4. Immediately before injecting the sample, remove the needle from the
stopper or septum, adjust the syringe down to the desired volume, and
inject the sample into the heated injection port on the GC. Note the
injection time on the strip chart and/or actuate the electronic integra-
tor.
5. Determine the retention time for each peak by dividing the distance of
the peak maximum from the injection point by the chart speed.
6. Repeat the analysis, and determine the retention times for each peak for
the second analysis. The retention times of successive injections should
agree within 0.5 seconds or within 1% of the mean of the adjusted
retention times, whichever is greater.
3.4.3 Analysis of Preliminary Survey Bag Samples - Follow the procedures described
in Subsection 5-3-1 for the analysis of bag samples. To assess the stability of
the gas sample in Tedlar bags, perform a second analysis after a time period equal-
ling the period between sample collection and the first analysis. If the concen-
tration of the sample collected in a Tedlar bag decreases by more than 1Q% between
the first and second analysis, then an accepted sampling method other than Tedlar
bags should be considered.
Perform a retention check on the bag sample by successively evacuating the bag
and refilling it with hydrocarbon-free air or nitrogen one or more times. Analyze -^
the bag contents for the target compound(s), allow the gas to sit in the bag f j
overnight, and reanalyze bag contents for the target compound(s). If any target V. J
compound is detected in the bag at a concentration greater than 5# of the original
concentration, then an accepted sampling method other than Tedlar bags should be
considered.
3.4.5 Analysts of Preliminary Survey Adsorption Tube Samples - Follow the proced-
ures described in Subsection 5-3-4 for the analysis of adsorption tube samples. A
minimum desorption efficiency of $0% must be obtained. If $0% desorption effici-
ency cannot be achieved using the referenced procedures from Table B in the Method
Highlights Section, then try longer desorption times, more vigorous desorption
techniques and/or other desorption solvents. If $0% desorption efficiency still
cannot be accomplished, then an accepted sampling method other than adsorption
tubes should be considered.
3.4.6 Interpretation of Preliminary Survey Eesults - To select the most suitable
sampling and analytical method for the final field test, the results of the prelim-
inary survey must be properly interpreted. The major points to consider are (1)
the sampling location, (2) the parameters of the process being tested, (3) the flue
gas moisture and temperature and the flue or duct static pressure, (4) stability of
the gas sample in bags, (5) the desorption efficiency of the target compounds from
adsorption tubes, and (6) the resolving capability, precision, accuracy, and speed
of the GC analysis. Thus, flue gas or duct parameters and components present
determine which sampling and analytical methodologies will be the most appropriate.
Sampling Location - The hazards associated with the sampling location will
influence the type of sampling methodology which can be used. In explosion risk
areas where use of pumps, heated probes, or a GC with a flame ionization detector
(FID) would be prohibited, the explosion risk area sampling procedure can be used
safely. Close attention must be paid to maintaining the proper sampling rate when
o
-------
Section No. 3.16.3
Date June 30, 1988
Page 19
using the evacuated canister as a vacuum source. It may be possible at certain
hazardous locations, to locate the collection device {bag or adsorption tube) in
the explosion risk area near the source and operate the sampling pump a safe dis-
tance away. Also, intrinsically safe personnel sampling pumps could be used in
certain hazardous areas provided the required sampling rate can be maintained.
Hazardous sampling locations may not be suitable for direct or dilution interface
sampling when using an electrically-heated probe and sample line.
Other physical factors concerning the sampling location will also influence
which sampling method is most suitable. These factors will be site-specific and
are beyond the scope of this Handbook.
Process Parameters - The particular process parameters pertaining to the
generation of the organic emissions and the effect the operation has on the
emission levels will influence which sampling technique will be most suitable. In
the case of a continuous process where emission levels are constant, each of the
Method 18 sampling techniques should be suitable with regard to the process
parameters. For processes operating in a batch or cyclic mode, the bag or adsorp-
tion tube integrated sampling techniques may be more suitable compared to the
interface techniques where grab samples are analyzed.
Flue Gas or Duct Conditions - The flue gas or duct moisture and temperature
will have a major influence on selecting the most suitable sampling technique.
High moisture will affect both bag samples and adsorption tube samples. For situa-
tions where moisture may be a problem, the interface techniques are recommended
provided the 5% criteria for consecutive injections, described in Subsections 5-3-2
and 5-3-^» can be met. Condensation in bag samples may result in the target organ-
ic compounds being absorbed into the condensate, or, at extremely high concen-
trations, being the condensate itself. The heated bag sampling technique may be
suitable provided on-site analysis is conducted when it is not practical to keep
the bags heated until analysis at the base laboratory. Condensation may also be
avoided by using a diluted bag sample collected by prefilling the bag with a known
quantity of hydrocarbon-free air or nitrogen prior to sampling and accurately
metering the gas sample into the bag during sampling.
Moisture reduces the adsorptive capacity of certain types of adsorbents (pri-
marily charcoal). . For sampling with adsorption tubes at sources with moisture
above 3%, a silica gel tube may be inserted in front of the primary adsorption
tube; otherwise, two or more adsorption tubes connected in series should be used.
The first tube becomes a sacrificial tube and should be positioned vertically
during sampling. A disadvantage of this approach is that the additional tubes will
also require analysis. Alternatively, a moisture knock-out jar can be used in
front of the adsorption tube. As varying amounts of the organic emissions will
also condense (the amount of each organic removed from the gas stream will largely
depend on the individual compound's volatility and solubility characteristics), the
liquid collected must be retained for analysis. Accurate quantitation of various
organics in the condensed liquid(s) may involve several steps and is generally
problematic.
The flue gas temperature may also dictate which sampling technique can be used
due to limitations of the sampling equipment.
Bag Sample Stability and Target Compound "Retention - If on-site analysis of
bag samples is not feasible and the samples are returned to the base laboratory for
analysis, then the stability of the gas sample in the bag will be a factor and
should be determined. While the stability of organics in bags has been demonstra-
-------
Section No. 3.16.3
Date June 30, 1988
Page 20
ted in numerous laboratory evaluations, an actual source sample could contain other
unknown components which may affect sample stability. If the preliminary survey
sample analysis indicates that the gas sample is not stable, then one of the alter-
native sampling procedures should be used. The check should be conducted at an
elevated temperature if the bag is to be heated during sampling.
The retention of the target compounds by the Tedlar bag should also be check-
ed. This check will indicate any sample loss not determined by the stability
check. If the retention of a target compound by the Tedlar bag is unacceptably
high, then the bag sampling technique is not suitable for that target compound and
one of the alternative sampling procedures should be used. Heating of the bag
during sampling and analysis may reduce the retention.
Adsorption Tube Desorption Efficiency - The desorption efficiency determined
for the adsorption tubes must be >50#. If >50% desorption efficiency cannot be
achieved with the referenced procedure, then more vigorous desorption techniques
and/or solvents should be evaluated. The desorption efficiency, as determined by
the procedures described in Subsection 5-1.6, will not indicate if the gas sample
matrix will affect the desorption of the target compounds. If acceptable
desorption efficiency cannot be achieved, then one of the alternative sampling
procedures should be used. Also, the adsorption efficiency must be greater than
90#. The breakthrough volume must not be exceeded.
Calibration Standards and GC Analysts - The availability of calibration stan-
dards may dictate which sampling technique can be used. The GC analysis may also
dictate which sampling technique will be the most suitable. For accurate analysis,
adequate resolution must be achieved between target compounds and between any
interfering compounds and target compounds. During preliminary survey sample
analysis, acceptable resolution may not be achievable on a gas sample but may be
accomplished with the adsorption tube sample, or vice versa. Thus, the sampling
technique which gives acceptable resolution during sample analysis must be select-
ed. In some situations where analysis of more than one target compound is requir-
ed, two or more analyses of the same sample under different GC conditions and/or
with different columns may be necessary to achieve adequate resolution.
Acceptable accuracy, as demonstrated by audit sample analysis, must also be
achieved for sample analysis by either gas or liquid injection. Again the sampling
technique that gives acceptable accuracy during sample analysis must be selected.
The sampling technique that gives acceptable precision, as demonstrated by consecu-
tive replicate injections, must be selected. Minimizing the analysis time is par-
ticularly important for the interface techniques. As discussed above, the preci-
sion limits may be hard to achieve with the interface techniques with a long analy-
sis time under variable or cyclic emission conditions.
3-5 Apparatus Check and Calibration
Figure 3-4 summarizes the pretest apparatus checks and calibration and can be
used as a pretest operations checklist. Figure 3-5 can serve as an equipment
packing list and status report form.
3.5-1 Probe - If a heated probe is required for the selected sampling procedure,
then the probe's heating system should be checked to see that it is operating prop-
erly. The probe should be cleaned internally by brushing first with tap water,
then with deionized distilled water, and finally, with acetone. Allow the probe to
air dry, then the probe should be heated and purged with air or nitrogen. The
o
o
o
-------
Section No. 3-16.3
Date June 30, 1988
Page 21
Date Calibrated by
Check Sampling Technique To Be Used
Direct Bag , Dilution Bag , Direct Interface , Dilution Interface ,
or Adsorption Tube .
Sampling Checks (Check only applicable methods)
Velocity and Water Vapor Content
Pitot tube dimension specifications checked? yes no (specification of Method
2, Handbook Section 3.1)
Differential pressure gauge pretest calibration acceptable? yes no N/A
(specifications of Method 2, Handbook Section 3.2)
Stack temperature sensor calibrated against a reference thermometer?* yes no
(within 5°F of reference thermometer)
Barometer pretest field barometer reading correct? yes no (within 2.5 nua
(0.1 in.) Hg of the mercury-in-glass barometer)
Wet bulb/dry bulb thermometers accuracy acceptable? yes no (within 1°F of
true value, manufacturer's specifications)
Method 4 sampling equipment acceptable?* yes no (Handbook Section 3.3, PRE
TEST SAMPLING CHECKS, Method 4, Figure 2.5)
Direct Bag
Pretest calibration of flowmeter acceptable? yes no (within 10 percent of
0.5 liter/min for single check)
For heated box system, pretest calibration of the temperature sensor in the box is
acceptable? yes no N/A (within 5 percent of reference value at
temperature of expected use)
Dilution Bag
Pretest calibration of flowmeter acceptable? yes no (within 3 percent of wet
test meter)
Pretest calibration factor of dry gas meter acceptable? yes no (within 2
percent of wet test meter)
*Most significant items/parameters to be checked.
(Continued)
Figure 3-4. Pretest sampling checks.
-------
Section No. 3.16.3
Date June 30, 1988
Page 22
Figure 3.4 (Continued)
Direct Interface and Dilution Interface
Pretest calibration of stack temperature sensor acceptable?* yes no (within
2°F of reference value)
Pretest calibration of probe and heated sample line temperature sensor acceptable?*
yes no (within 2°F of reference value)
For dilution interface only, pretest calibration of dilution system acceptable?*
yes no (within 10 percent of expected dilution factor)
Pretest calibration of gas chromatograph acceptable?* yes no (specifications
shown in POSTSAMPLING OPERATIONS CHECKLIST, Figure 5.10)
Adsorption Tubes
Pretest calibration of limiting orifice acceptable?* yes no (compared to
bubble meter)
*Most significant items/parameters to be checked.
o
o
'.- \ V
I; ' '
-------
Section No. 3.16.3
Date June 30, 1988
Page 23
Apparatus check
Moisture Determination
V Bulb/D Bulb
Checked
Barometer
Calibrated*
Method 4
Probe, heated &
leak checked
Impingers
Meter system
calibrated*
Velocity Determination
Pitot Tube
Number
Length
Pressure Gauge
Manometer
Other
Stack Thermometer
Calibrated
Bag Sampling
Probe Liner
S steel
Glass
Teflon tube
Length
Meter System
Flowmeter*
Pump
Evacuated can
Charcoa I tube
Sample line
Tedlar Bags
Number
Blank checked
Heated Box
Number
Heat checked
Acceptable
Yes
No
Quantity
Required
Ready
Yes
No
Loaded and Packed
Yes
No
*Most significant items/parameters to be checked.
Figure 3-5- Pretest preparations.
-------
Section No. 3-16.3
Date June 30, 1988
Page 24
Figure 3.5 (Continued)
o
Apparatus check
Bag Sampling (continued)
Dilution
Inert gas
Meter
Gas Chromatograph
On-site
(check below)
N/A
Direct or Dilution
Interface
Probe Liner
Glass
S steel
Teflon
Heated Line
Length
Heat checked
Temperature Sensors
Stack
Probe
-Calibrated*
Sample Pump
Dilution System
Dilution pumps
Flowmeters
Dilution gas
Heated box
Dilution factor
checked*
Gas Chromatograph
(shown below)
Adsorption Tube
Probe
Heated
Checked
Nonheated
Glass
S steel
Filter
Acceptable
Yes
No
Quantity
Required
Ready
Yes
No
Loaded and Packed
Yes
No
O
*Most significant items/parameters to be checked.
O
-------
Section No. 3.16.3
Date June 30, 1988
Page 25
Figure 3.5 (Continued)
Apparatus Check
Adsorption Tube (continued)
Sample Line
Type
Length
Checked*
Pump and Meters
Pump
Orifice
Calibrated*
Eotameter
Calibrated*
Timer
Adsorption Tubes
Type
Gas Chromatograph
Cylinder Standards
Analyte
PPM
PPM
Regulators
Bags
Size
Dilution system
Calibrated*
Diluent gas
Bag Standards
Analyte
Syringes
Iwptnger/hot
plate assembly
Gas meter
N2 gas
Regulator
Bags
Size
Other Gases
Fuel
Carrier
Zero
Columns
Type
Acceptable
Yes
No
Quantity
Required
Ready
Yes
No
Loaded and Packed
Yes
No
*Most significant items/parameters to be checked.
-------
Section No. 3.16.3
Date June 30, 1988
Page 26
Figure 3.5 (Continued)
Apparatus Check
Gas Chromatograph
Type
Temp /con oven
Bulk Samples
Bottles
Type
Size
Acceptable
Yes
No
Quantity
Required
Ready
Yes
No
Loaded and Packed
Yes
No
o
*Most significant items/parameters to be checked.
O
o
-------
Section No. 3-16.3
Date June 30, 1988
Page 27
probe should be sealed at the inlet end and checked for leaks by applying a vacuum
of 10 in. H20. See Subsection 1.0 for the probe leak check procedure. The probe
is considered leakfree under these conditions if no loss of vacuum is seen after
one minute. Any leaks should be corrected or the probe should be rejected. If the
probe has an external sheath, the integrity of the seal between the sheath and the
probe liner should be checked to ensure ambient air does not dilute the gas sample.
3. 5 -2 Teflon Tubing - Prepare enough sections of tubing for connecting the probe
to bag or tube samples.
3.5.3 Quick Connects or Equivalent - The quick connects, or their equivalents,
should be new or cleaned according to the manufacturer's recommendations. Leak
check the quick connects as described in Subsection 1.0.
3.5.4 Barometer - The field barometer should be compared with the mercury-in-glass
barometer or with a National Weather Service Station reading prior to each field
test.
3.5.5 Met Bulb/Dry Bulb Thermometers - For sources with stack temperatures below
59° C where wet bulb/dry bulb thermometers will be used to determine stack gas
moisture content, the thermometers should be compared with the mercury-in-glass
thermometer at room temperature prior to each field trip.
3.5.6 Method 4 Equipment - Where Method 4 will be used to determine stack gas
moisture content, prepare the equipment for sampling following the procedures
described in Section 3-3.3 of this Handbook.
3.5.7 S-type Pitot Tube and Differential Pressure Gauge - Prepare the S-type pi tot
tube and the differential pressure gauge for sampling following the procedures de-
scribed in Section 3-1.3 of this Handbook.
3.5.8 Sampling Pump - Check the sampling pump for delivery rate and leaks before
going to the field as follows: Attach a 0 to 5 liter/minute rotameter, to the
outlet of the pump and turn on the pump. Check the flow rate indicated by the
rotameter. Reject or repair the pump if the flow rate is not at least 1 liter/mi-
nute. If the flow is adequate, then conduct a leak check by plugging the inlet of
the pump. If the pump is leak free then the rotameter should eventually indicate
no flow. Repair or replace the pump if a leak is indicated.
3.5.9 Tedlar Bags - Prepare new Tedlar bags for sampling by leak checking the bags
before going to the field. The bags should also be checked for contamination by
filling with hydrocarbon- free air or nitrogen during the leak check. The bags are
checked as follows: Connect a water manometer, or equivalent, using a tee con-
nector between the check valve quick connect on the bag and a pressure source (or
hydrocarbon-free air or nitrogen for conducting the contamination check) . Pres-
surize the bag to 5 to 10 cm (2 to 4 in.) H20 and disconnect the quick connect.
Loss of pressure over a 10 minute period indicates a leak. Alternatively, leave
the bag pressurized overnight; a deflated bag the following day is indicative of a
leak. Reject or repair any bags with leaks. After the hydrocarbon- free air or
nitrogen has remained in the bag for 24 hours, analyze the bag contents using a GC
with a flame ionization detector on the most sensitive setting. The bag should be
rejected if any organic compounds are detected that may interfere with the analysis
of any of the target compound (s) .
-------
Section No. 3.16.3
Date June 30, 1988
Page 28
3.5.10 Etgtd Leak-Proof Containers - The rigid containers used to contain the
Tedlar bags during sampling should be checked for leaks prior to going to the
field. The container should be leak checked with the bag in place as follows:
Connect a water manometer, or equivalent, using a tee connector between a pressure
source and the container outlet. Pressurize the container to 5 to 10 cm (2 to 4
in.) Hg. Any loss of pressure after 10 minutes indicates a leak. Reject or repair
the rigid container if a leak is indicated.
3.5-11 Direct Pump Sampling System - If the direct pump sampling system is select-
ed, then the system should be assembled and leak checked prior to going to the
field as follows: Assemble the system (see Figure 4.5). Attach a vacuum line and
a rotameter to the inlet quick connect. Plug the probe inlet and turn on the
vacuum pump. If the system is leakfree up to the pump, the rotameter should even-
tually indicate no flow. An alternate procedure to leak check the system up to the
male inlet check valve quick connect is as follows: Connect a water manometer, or
equivalent, using a tee connector between a pressure source and the inlet end of
the probe. Pressurize the system to 5 to 10 cm (2 to 4 in.) Hg. Any loss of
pressure after 30 seconds indicates a leak. Reject or repair the sampling system
if a leak is indicated. Check to see if the pump is contaminating the sampling
system by filling a second contamination-free Tedlar bag with hydrocarbon-free air
or nitrogen, and with the system assembled pull the hydrocarbon-free air or nitr-
ogen from the second Tedlar bag into the first Tedlar bag using the pump. Analyze
the first bag contents using a GC with a flame ionization detector on the most
sensitive setting. The pump should be rejected or repaired, cleaned, and checked
again if any organic compounds are detected that may interfere with the analysis of
any of the target compound(s).
3.5.12 Needle Valve and Rotameter - Prior to each field trip or at sign of erratic
behavior, the flow control valve and the rotameter should be cleaned according to
the maintenance procedure recommended by the manufacturer.
3.5-13 Teflon Probe - For bag sampling in an explosion risk area, prepare a new
Teflon probe or clean a used Teflon probe following the procedure described in
Subsection 3•5•1 - Leak check the Teflon probe as follows: Attach a mercury manome-
ter, with a tee connector, and a vacuum pump to the outlet of the probe. Plug the
inlet end of the probe and apply a vacuum of 10 in. H20. The probe is considered
leak free under these conditions if no loss of vacuum is seen after one minute.
Any leaks should be corrected or the probe should be rejected.
3-5-14 Explosion Risk Area Sampling System - The explosion risk area sampling
system should be leak checked as follows: Evacuate the steel drum. Assemble the
system (see Figure 4.6), with the pinch clamp open, the sample bag leak checked and
evacuated, and directional needle valve closed. Attach a mercury manometer to the
inlet of the Teflon probe. Open the needle valve. The rotameter should eventually
indicate no flow. Once there is no flow, note the manometer reading. The system
is considered leak free under these conditions if no loss of vacuum is seen after
one minute. Any leaks should be corrected or the system should be rejected. It is
recommended that an explosion-proof pump be used in the explosion risk area or a
regular pump be used outside the risk area. Follow the procedures described for
these pumps.
3-5-15 Heated Bag Sample Container and Sample Lines - If other modified bag sampl-
ing techniques are selected due to condensation observed during sampling, heated
o
o
o
-------
Section No. 3-16.3
Date June 30, 1988
Page 29
bag sample containers and sample lines will be required. The heating systems of
this equipment should be checked prior to going to the field to see that they are
operating properly. The sample lines should be cleaned following the procedure
described for the sampling probe in Subsection 3-5.1. The heated sampling system
should be assembled and leak checked prior to going to the field as follows:
Assemble the system (see Figure 4.5). Attach a vacuum pump and a rotameter to the
inlet quick connect. Plug the probe inlet and turn on the vacuum pump. If the
system is leak free, the rotameter should eventually indicate no flow. An alter-
nate procedure to leak check the system up to the female outlet check valve quick
connect on the bag container is as follows: Connect a water manometer, or equiva-
lent, using a tee connector between a pressure source and the inlet end of the
probe. Pressurize the system to 5 to 10 cm (2 to 4 in.) Hg. Any loss of pressure
after 30 seconds indicates a leak. Reject or repair the sampling system if a leak
is indicated.
3.5.16 Direct Interface Sampling System - The heating system of the sampling
probe should be checked prior to going to the field if heating is required to
maintain the gas sample above the duct temperature and/or to prevent condensation.
The probe should also be cleaned and leak checked following the procedures describ-
ed in Subsection 3«5-l. If the probe has an external sheath, the integrity of the
seal between the sheath and the probe liner should be checked to ensure ambient air
does not dilute the gas sample. The sample line should be cleaned following the
procedure described for the sampling probe in Subsection 3-5.1. The heating system
of the sample line should be checked before going to the field to see that it is
operating properly. The direct interface sampling system should be assembled and
leak checked prior to going to the field as follows: Assemble the system (see
Figure 4.5). Switch the gas sampling valve to the inject position, and plug the
outlet from the sample valve. Connect a water manometer, or equivalent, using a
tee connector between a pressure source and the inlet end of the probe. Pressurize
the system to 5 to 10 cm (2 to 4 in.) Hg. Any loss of pressure after 30 seconds
indicates a leak. Reject or repair the sampling system if a leak is indicated.
3.5-17 Dilution Interface Sampling System - The equipment required for dilution
interface sampling is the same as required for direct interface sampling, with the
addition of a heated dilution system and a larger heated sample pump. The heating
systems should be checked to see that they are operating properly. Prior to each
field trip or at sign of erratic behavior, all flowmeters should be cleaned accord-
ing to the maintenance procedure recommended by the manufacturer. The flowmeters
should also be calibrated following the procedures described in Subsection 2.1-3-
The dilution interface sampling system should also be checked for leaks as follows:
Assemble the system (see Figure 4.6). Connect a water manometer, or equivalent,
using a tee connector between a pressure source and the inlet end of the probe.
Plug the three outlet vents to the charcoal adsorbers and the outlet of the two
flowmeters. Pressurize the system to 5 to 10 cm (2 to 4 in.) Hg. Any loss of
pressure after 30 seconds indicates a leak. Reject or repair the system if a leak
is indicated. It is advisable to verify the operation of the dilution system prior
to going to the field following the procedures described in Subsections 4.3-7 and
5.3.3.
3.5.18 Gas Chromatography System - Refer to Table C in the Method Highlights
Section to ensure that the proper detector has been selected for the target organic
compounds. Prior to taking the gas Chromatography system to the field, check that
all systems are operating properly. Consult the operator's manual for procedures
-------
o
Section No. 3-16.3
Date June 30, 1988
Page 30
to verify that the equipment is operating properly. Check to see that all cylinder
gas regulators, connections, and tubing are functioning properly and are leak free.
This is particularly important when using hydrogen and oxygen. Consult with the
regulator manufacturer for procedures on checking pressure regulators. Connections
and tubing can be checked for leaks by pressurizing with the gas and wetting with a
soap solution or other commercially available solutions. Any bubbles forming on
the connections indicate a leak. Tighten or replace any leaking connections. An
alternative leak check procedure for carrier gas is as follows: Plug the outlet of
the analytical column. Pressurize the tubing and connections. Turn off the cylin-
der valve and note pressure on the regulator gauge and, if equipped, the GC pres-
sure gauge. Any loss of pressure indicates a leak. Locate the leak using a soap
solution or work backwards through the carrier gas flow path disconnecting each
component and plugging the carrier gas flow until the leak is located, a similar
check should be made of the gas sampling valve, sample loop, and connections.
It is advisable to take to the field an adequate supply of spare parts, sep-
tums, different size sample loops, extra analytical columns, and other related
equipment that may fail or deteriorate during the emission test. The generation of
response factors for each target organic compound, relative to a single organic
compound, should be confirmed in the laboratory prior to going to the field. The
confirmation procedure involving preparation and analysis of calibration standards
containing multiple organic compounds is described in Section 5.1.6.
3.6 Reagents and Equipment
The following reagents and equipment may be required 'to conduct the emission
test depending on the sampling method selected. These materials are generally ac-
quired from commercial vendors. Certification of purity and/or analysis should be
obtained for adsorption tubes, calibration and zero gases, and liquid organic com-
pounds .
3.6.1 Charcoal Adsorber - Check to see that the supply of charcoal adsorbent is
sufficient to last for the entire field test period.
3.6.2 Adsorption Tubes - If adsorption tube sampling is to be conducted, check to
see that the proper cype of tube has been obtained for collecting the target or-
ganic compounds. Refer to Table B in the Methods Highlights Section to determine
the proper p^sorption material. Check to see that the supply of adsorption tubes
is suffi^ent to conduct the emission test, including field blanks and for desorp-
ti™ efficiency determinations.
3.6.4 GC Carrier Gas - Check the GC operator's manual and the GC column manufac-
turer to see that the GC carrier gas type and grade are compatible with the GC and
the column. Check to see that the supply of carrier gas is sufficient to last the
entire field test period.
3.6.4 Auxiliary GC Gases - Check to see if the proper type and grade of auxiliary
gases required by the GC detector have been obtained. Consult with the GC detector
manufacturer to determine the proper type and grade of auxiliary gases required.
Check to see that the supply of auxiliary gases is sufficient to last the entire
field test period. x—x
3.6.5 Calibration Gases - Check to see if the correct calibration gases in the re- V_x
quired range have been obtained. If available, commercial cylinder gases may be
o
-------
Section No. 3.16.3
Date June 30, 1988
Page 31
used if their concentrations have been certified by direct analysis; cylinder gases
with tighter tolerances on their concentrations are preferred. Check to see that
the supply of calibration gases is sufficient to last the entire field test period.
3.6.6 Calibration Gas Dilution System - Prior to each field trip or at the sign of
erratic behavior, any flow control valves or rotameters used in the dilution system
should be cleaned according to the maintenance procedure recommended by the manu-
facturer. The rotameters or other metering devices used with a single-stage or
two-stage dilution system should be calibrated prior to going to the field follow-
,ing the procedures described in Subsection 2.2. It is advisable to check the
dilution ratio of the dilution system prior to going to the field following the
procedures described in Subsections 4.3.7 and 5-3-3-
3.6.7 Zero Gas - Check to see that the zero gas meets the requirements for being
hydrocarbon-free (less than 0.1 ppmv of organic material as propane or carbon equi-
valent) . Check to see that the supply of zero gas is sufficient to last the entire
field test period.
3.6.8 Audit Gases - Check to see that the required audit gases in the proper range
have been acquired. Consult Table A in the Method Highlights Section for audit
gases available from the EPA for the target organic compounds. The availability
and ranges of audit gases can be determined by contacting:
Environmental Protection Agency
Environmental Monitoring Systems Laboratory
Quality Assurance Division (MD-77B)
Research Triangle Park, North Carolina 27711
Attention: Audit Cylinder Gas Coordinator
For audit gases obtained from a commercial gas manufacturer, check that the manu-
facturer has (1) certified the gas in a manner similar to the procedure described
in 40 CFR Part 61, Appendix B, Method 106, Section 5.2.3.1 and (2) obtained an
independent analysis of the audit cylinder that verifies that the audit gas concen-
tration is within 5% of the manufacturer's stated concentration.
3.6.9 Organic Compounds for Preparing Gaseous Standards - If gaseous standards are
to be prepared in the field, check to see if the organic compounds to be used are
at least 99-9# pure or, if less than 99-9#» of known purity necessary to calculate
the gaseous standard concentration. Record the manufacturer's lot number for each
standard compound.
3.6.10 Equipment for Preparing Gaseous Standards by Liquid or Gas Infection-
Confirm that the Tedlar bags to contain the gaseous standards have been leak check-
ed following the procedures described in Subsection 3-5-9- Check to see that the
syringes selected are gas-tight, cover the range needed (1.0- to 10-microliters for
liquids and 0.5 ml for gases), and are accurate to within i%. Confirm that the dry
gas meter and temperature gauge have been calibrated following the procedures
described in Subsection 2.0. Clean the midget impinger assembly with detergent and
tap water, and then rinse with deionized distilled water. Check the system for
leaks as follows: Assemble the appropriate system for preparing standards (see
Figure 5.5 for gaseous materials or Figure 5.6 for liquid materials). Fit the
injection port with a new septum. Fill the Tedlar bag and pressurize the system
to 5 to 10 cm (2 to 4 in.) Hg. Any loss of pressure after 10 minutes indicates a
-------
Section No. 3.16.3
Date June 30, 1988
Page 32 S~\
leak. Reject or repair the system if a leak is indicated.
3.7 Packing Equipment for Shipment
The packing techniques described in this section are not requirements, but are
suggestions based on previous field experience. The type of packaging for equip-
ment going to the field depends on the mode of transportation. Typically, packing
equipment for transport by a common carrier will require the greatest degree of
effort to ensure the equipment arrives on-site in its original condition. When
possible, delicate equipment should be packed in the original shipping containers.
For convenience, label all containers with the contents for easy identification in
the field. The most common mode of packing will be in a van or trailer, where the
equipment will remain during transport. More sophisticated test firms have trai-
lers or trucks dedicated to the type of sampling being conducted. These units are
often designed to allow the test equipment and instruments to remain set up during
transport. This approach minimizes the time and effort required to set up before
and breakdown after a test. A dedicated test vehicle provides a working environ-
ment that greatly enhances the quality of work that can be performed.
3.7-1 Probe - Pack the probe in a rigid case protected by polyurethane foam, poly-
ethylene bubble-pack, or other suitable packing material. Seal the inlet and
outlet of the probe with tape or other suitable material. Protect any protruding
gloss ends from breakage by insertion into rigid plastic pipe lined with foam or
other packing material.
3.7-2 Teflon Tubing, Sample Lines, and Vacuum Lines - All tubing, sample lines,
and vacuum lines should be coiled and secured with tape. Coils should be large
enough not to crimp tubing or excessively strain the heat sheath. Seal all open-
ings with tape.
3-7-3 Quick Connects, Flow Control Valves and other Connectors - All connectors,
valves, and other small parts should be packed in small parts cabinets, trays with
divided compartments, or storage chests with labeled drawers to provide quick and
easy access to the desired part.
3.7.4 Barometer - The field barometer should be packed in a rigid container,
securely mounted in rigid foam. The barometer case should be packed in a larger
box designated to contain delicate or fragile equipment.
3.7-5 Thermometers and Thermocouple Readouts - Thermometers and thermocouple read-
outs should be packed in the original carrying case, if possible. Glass thermome-
ters should be packed in a rigid tube to prevent breakage. These items, in their
smaller packing, should also be packed in a larger box designated to contain deli-
cate or fragile equipment.
3«7-6 Method 4 Equipment - Method 4 equipment should be packed following the pro-
cedures recommended in Section 3.3.3 of this Handbook.
3.7.7 S-type Pitot Tube and Differential Pressure Gauge - The S-type pitot tube,
when not mounted on the sampling probe, should packed in a rigid case and wrapped
with polyurethane foam, polyethylene bubble-pack, or other suitable type of packing
material. Seal all openings with tape or other, suitable material. The differen-
tial pressure gauge, if not part of a meter box, should be mounted in a rigid
o
o
-------
Section No. 3.16.3
Date June 30, 1988
Page 33
housing. The gauge should be wrapped with polyurethane foam, polyethylene bubble-
pack, or other suitable material, and packed in a larger box designated for deli-
cate and fragile equipment.
3-7-8 Glassware - All glassware should be packed in the original shipping contain-
ers, if available, and stored together in a larger rigid container marked "Fragile!
Glass." Otherwise, wrap the glassware with polyurethane foam, polyethylene bubble-
pack, or other suitable material, and pack in a rigid foam-lined container marked
"Fragile! Glass."
3.7.9 Tedlar Bags - Preferably, transport the Tedlar bags to the field in individ-
ual rigid containers used for sampling. If this is not possible, pack the bags,
individually, in corrugated cardboard boxes with the connectors secured such that
they do not contact and puncture the bags.
3-7-10 Sampling Pumps - Sampling pumps, if not mounted in a rigid housing suitable
for transport, should be packed in a rigid foam-lined container.
3.7.H Dilution Interface System - The dilution interface system should be built
into in a rigid container suitable for shipment. • :
3-7-12 Gas Chromatograph System - The gas chromatograph and ancillary systems
should be packed in the original shipping container for transport. Although it is
not recommended, the GC can be transported with out additional! packaging in a van
or trailer provided the GC is secured properly against movement and other equipment
is not packed in a manner where it could fall on the instrument. For transport in
dedicated test vehicles, the instruments should be mounted in shock absorbing
devices. All gas lines and analytical columns should be capped to prevent con-'
tamination and/or oxidation during shipment.
3-7-13 Gas Cylinders - All gas cylinders should be transported with their protec-
tive cylinder heads securely attached. The cylinders should be secured horizontal-
ly so that they do not roll together or vertically in a specially designed cylinder
rack. Be aware of and adhere to all Federal, State, and local regulations involv-
ing the transport of compressed and flammable gases, particularly through tunnels.
3.7.14 Liquid Organic Compounds - Liquid organic compounds should be shipped with
the container top sealed with electricians tape and stored in a sealed plastic bag.
Packed each container in its original shipping box, if available. Otherwise, wrap
each container individually with polyurethane foam, polyethylene bubble-pack, or
other suitable material and place in a box designated for chemicals.
3.7.15 Dry Gas Meters - Dry gas meters not housed in a rigid meter box suitable
for transport should be wrapped with polyurethane foam, polyethylene bubble-pack,
or other suitable material, and packed in a larger box designated for delicate or
fragile equipment.
-------
Section No.- 3-16.3
Date June 30, 1988
Page 34
Table 3.1. ACTIVITY MATRIX FOR PRELIMINARY SURVEY SAMPLING AND ANALYSIS
Characteristic
Apparatus Check
Barometer
Wet bulb/dry bulb
thermometers
Method 4 equipment
S-type pitot tube
and differential
pressure
Probe
Teflon tubing
Quick connects
Glass flasks
High-vacuum pump
Tedlar or alumi-
nized Mylar bags
Acceptance limits
Within 2.5 mm
(0.1 in.) Hg of
mercury-in-glass
barometer
Within 1°C (2°F) of
a mercury-in-glass
thermometer
See Section 3.3.3
of this Handbook
See Section 3.1.3
of this Handbook
1. Clean; glass
liner, stainless
steel, or Teflon
inert to organics
2. Heating properly
if equipped with
heating system
3. Leak free
New and unused
New or clean
Clean
Vacuum of 75 n™
(3 in.) Hg absolute
Leak free; no
loss of pressure
after 30 seconds
Frequency and method
of measurement
Before each field trip
As above
Same as Section 3.3«3
Same as Section 3- 1-3
Before each field trip
following the proced-
ures described in Sub-
section 3«5-l
As above
As above
As above
As above
As above
As above
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20
Action if
requirements
are not met
Repair or replace
Replace
Same as Section
3.3.3
Same as Section
3.1.3
Repeat cleaning
Repair or replace
As above
Obtain new tubing
Clean according
to manufacturer's
recommendation
Repeat cleaning
of flasks
Repair or replace
As above
o
o
(Continued)
-------
Section No. 3-16.3
Date June 30, 1988
Page 35
Table 3.1 (Continued)
Charac teris tic
Apparatus Check
Rigid containers
Direct pump
sampling system
Needle valve and
rotameter
Adsorption Tube
Procedure
Adsorption tubes
Personnel sampling
pump
Extraction solvent
Teflon tubing
On-site Measure-
ments and Sampling
Wet bulb/dry bulb
measurement
Acceptance limits
Leak free; no
loss of pressure
after 30 seconds
Leak free; no
loss of pressure
after 30 seconds
Clean
Proper type of
adsorption material
Calibrated
Proper type of
extraction solvent
New and unused
1. Wet bulb wick
moistened
2. Wet bulb temper-
ature stabilized
3. Record wet bulb
and dry bulb
temperature
Frequency and method
of measurement
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20
As above
Prior to each trip or
at the sign of erratic
behavior
Before each field trip
As above
Prior to extraction
of tubes for analysis
Before each field trip
Prior to each
measurement
During measurement
Immediately after wet
bulb temperature
stabilizes
Action if
requirements
are not met
Repair or replace
As above
Clean following
manufacturer1 s
recommendations
Replace with
proper type
Repair or replace
Replace with
proper type
Obtain new tubing
Moisten
Allow to
stabilize
Repeat
measurement
(Continued)
-------
Section No. 3.16.3
Date June 30, 1988
Page 36
Table 3.1 (Continued)
o
Characteristic
On-site Measure-
ments and Sampling
Evacuated glass
flask sampling
Purged glass
flask sampling
Acceptance limits
1. Flask evacuated
to pump capacity
2. Assemble; no
leakage
3« System purged up
to flask inlet
4. Stopcocks closed
and taped; flask
labeled
5- Flue gas tempera-
ture and static
pressure determined
1. Assemble; no
leakage
3. Entire system
purged for 2 minutes
3. Stopcocks closed
and taped; flask
labeled
4. Flue gas tempera-
ture and static
pressure determined
Frequency and method
of measurement
Prior to sample
collection
Before sample col-
lection, visually and
physically inspect
all connections
Immediately prior to
sampling
Immediately after
sampling
Immediately after
sampling
Before sample col-
lection, visually and
physically inspect
all connections
Immediately prior to
sampling
Immediately after
sampling
Immediately after
sampling
Action if
requirements
are not met
Evacuate flask
Check for leaks;
repair system;
repeat test
Purge system up
to flask inlet
Close and tape
stopcock; label
flask
Determine flue
gas temperature
and static
pressure
O
Check for leaks;
repair system;
repeat test
Purge entire
system for 2
minutes
Close and tape
stopcock; label
flask
Determine flue
gas temperature
and static
pressure
(Continued)
O
-------
Section No. 3-16.3
Date June 30, 1988
Page 37
Table 3.1 (Continued)
Characteristic
On-site Measure-
ments and Sampling
Flexible bag sam-
pling procedure
Adsorption tube
sampling procedure
Preliminary Survey
Sample Analysis
Calibration
standards
(Continued)
Acceptance limits
1. Assemble using
Figure 4.4; no
leakage
2. Flow rate set to
0.5 1pm; purge sy-
stem up to bag inlet
3. Flue gas tempera-
ture and static
pressure determined
4. Bag labeled and
protected from
sunlight
1. Assemble using
Figure 4.9. no
leakage
2. Tubes capped,
labeled and stored
3- Flue gas tempera-
ture and static
pressure determined
1. Minimum of three
standards prepared
for each analyte
2. Sufficient peak
resolution achieved
(valley height <25#
of the sum of the 2
peak heights)
Frequency and method
of measurement
Before sample col-
lection, visually and
physically inspect
all connections
Immediately prior to
sampling
Immediately after
sampling
Immediately after
sampling
Before sample col-
lection, visually and
physically inspect
all connections
Immediately after to
sampling
Immediately after
sampling
Prior to sample
analysis
During multiple
component standard
analysis
Action if
requirements
are not met
Check for leaks;
repair system;
repeat test
Set flow rate
Purge system up
to flask inlet
Determine flue gas
temperature and
static pressure
Label bag and
protect from
sunlight-
Check for leaks;
repair system;
repeat test
Cap, label and
store tubes
Determine flue
gas temperature
and static press.
Prepare three
standards for
each analyte
Vary GC operating
conditions and/or
change column
type
-------
Table 3.1 (Continued)
Section No. 3.16.3
Date June 30, 1988
Page 38
o
Characteristic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Preliminary Survey
Sample Analysis
Calibration
standards
3. Response for
consecutive repli-
cate injections of
each standard agree
within 5% of their
average response
4. Calibration curve
generated
5. Audit sample
(optional) analysis
results within iO%
of true value
During calibration
standard analysis
Repeat injections
After calibration
standard analysis
As above
Perform regres-
sion analysis and
plot curve
Repeat audit;
remake and
reanalyze
standards
O
Glass flask sample
analysis
(Continued)
1. Condensation in
sample flask
2. Flask not
pressurized
3. Condensation in
pressurized flask
after 10 minute
equilibration
4. Adequate resolu-
tion between peaks
achieved for peaks
>5% of total area
5. Retention times
of consecutive in-
jections determined
and agree within 0.5
seconds or 1%
Before sample analysis
As above
As above
During sample analysis
After sample analysis
Heat flask to
flue gas or duct
temperature
Pressurize flask
Heat flask to
vaporize conden-
sate; if flask
already heated,
release pressure
and repressurize
Vary GC operating
conditions and/or
change column
type
Repeat analysis
O
-------
Section No. 3-16.3
Date June 30, 1988
Page 39
Table 3.1 (Continued)
Characteristic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Preliminary Survey
Sample Analysis
Flexible bag
samples
1. Response for
consecutive repli-
cate injections of
each sample agree
within 5% of their
average response
2. Stability of bag
samples acceptable
(second analysis
conducted an equal
number of days cor-
responding to the
the time elapsed
between sample col-
lection and first
analysis within 10%)
During sample analysis
After second analysis
Repeat analysis;
diagnose GC
problem
Consider one
of the alternate
sampling methods
Adsorption tube
samples
1. Samples desorbed
for period specified
in referenced method
2. Response for
consecutive repli-
cate injections of
each sample agree
within 5% of their
average response
3. Desorption effi-
ciency >50#
Before sample analysis
During sample analysis
After sample analysis
Check referenced
method; desorb
for specified
period
Repeat analysis;
diagnose GC
problem
Evaluate more
vigorous desorp-
tion techniques;
Consider one of
the alternative
sampling methods
-------
Section No. 3.16.3
Date June 30, 1988
Page 40
Table 3.2. ACTIVITY MATRIX FOR PRESAMPLINQ PREPARATION
Characteristic
Apparatus Check
Barometer
Wet bulb/dry bulb
thermometers
Method 4 equipment
S-type pitot tube
and differential
pressure
Probe
Teflon tubing
Quick connects
Sampling pump
Tedlar bags
Rigid containers
Acceptance limits
Within 2.5 mm
(0.1 in.) Hg of
mercury-in-glass
barometer
Within 1°C (2°F) of
a mercury-in-glass
thermometer
See Section 3.3.3
of this Handbook
See Section 3.1.3
of this Handbook
1. Clean; glass
liner, stainless
steel, or Teflon
inert to organics
2. Heating properly
if equipped with
heating system
3- Leak free
New and unused
New or clean
Leak free; adequate
delivery (>^ 1 Lpm)
Leak free; no
loss of pressure
after 10 minutes
Leak free; no
loss of pressure
after 30 seconds
Frequency and method
of measurement
Before each field trip
As above
Same as Section 3.3.3
Same as Section 3- 1.3
Prior to each trip
follow the cleaning
procedure described
in Subsection 3«5«1
Prior to each trip
As above
As above
As above
Prior to each trip
check with a rotameter
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20
Action if
requirements
are not met
Repair or replace
Replace
Same as Section
3.3.3
Same as Section
3-1.3
Repeat cleaning
Repair or replace
As above
Obtain new tubing
Clean according
to manufacturer's
recommendation
Repair or replace
As above
As above
o
o
o
(Continued)
-------
Section No. 3.16.3
Date June 30, 1988
Page 41
Table 3.2 (Continued)
Characteristic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Apparatus Check
Direct pump
sampling system
Leakfree; no
loss of pressure
after 30 seconds
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.)
As above
Needle valve and
rotameter
Clean
Prior to each trip or
at the sign of erratic
behavior
Clean following
manufacturer's
recommendations
Explosion risk
area sampling
system
Leakfree (no vacuum
loss after 1 minute)
Prior to each trip
Repair or replace
Heated bag
sampling container
1. Leakfree; no
loss of pressure
after 30 seconds
2. Heating properly
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20
As above
As above
As above
Direct interface
sampling system
1. Heating properly
2. Leakfree (no
flow at rotameter
with probe plugged)
As above
As above
As above
As above
Dilution interface
sampling system
1. Heating properly
2. Flowmeters cali-
brated
3. Leakfree; no
loss of pressure
after 30 seconds
As above
Calibrate prior to
each test against a
bubble meter or
spirometer
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H.,0
As above
Calibrate
Gas chromatograph
equipment
Leakfree, opera-
tional, and suffi-
cient spare parts
for the duration of
the field test
Prior to field test
check system for
leaks, access opera-
tional condition, and
inventory spare parts
Consult the
operator's
manual
(Continued)
-------
Section No. 3.16.3
Date June 30, 1988
Page 42
o
Table 3.2 (Continued)
Characteristic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Reagents and
Equipment
Charcoal adsorber
Sufficient supply
Check supply prior to
each field test
Procure more
adsorbent
Adsorption tubes
Proper adsorbent,
tube size, and
quantity for test
Prior to field test
refer to Method
Highlights Section and
preliminary survey
results
Procure proper
adsorbent, tube
size, and
quantity
Gas chromatograph
carrier gas
Carrier gas compat-
ible to GC and
column; sufficient
quantity for test
Prior to field test
refer to operator's
manual or consult
with manufacturer
Procure compat-
ible carrier gas
in sufficient
quantity
O
Auxiliary GC gases
Proper type and
grade for GC detec-
tor; sufficient
quantity for test
Prior to field test
refer to operator's
manual or consult
with manufacturer
Procure proper
type and grade
of gases in suf-
ficient quantity
Calibration gases
Proper component(s)
and range; suffi-
cient quantity for
any on-site calibra-
tions
Prior to field test
refer to the prelim-
inary survey results
Obtain gases with
the proper compo-
nents in the
necessary range
and quantity
Calibration gas
dilution system
1. Rotameters clean
and calibrated
2. Dilution ratio
known (optional)
Prior to field test
examine and calibrate
following procedures
in Subsection 2.2
Prior to field test
establish the ratio
following the proced-
ures in Section 5-0
Clean and
calibrate
Check dilution
ratio prior to
analysis
(required)
Zero gas
Hydrocarbon-free
(<0.1 ppmv as pro-
pane or carbon equi-
valent) ; sufficient
supply for test
Analyze or consult
manufacturer
Procure hydro-
carbon-free gas
in sufficient
quantity for test
O
(Continued)
-------
Section No. 3.1613
Date June 30, 1988
Page 43
Table 3.2 (Continued)
Characteristic
Audit gases
Organic compounds
for preparing
gaseous standards
Equipment for
preparing gaseous
standards
Packing Equip-
ment for Shipment
Probe
Teflon tubing,
sampling lines,
and vacuum lines
Quick connects,
flow control
valves, and other
connectors
Barometer
Thermometers and
thermocouple read-
outs
Method 4 equipment
S-type pitot tube
and differential
pressure gauge
Acceptance limits
Required audit gases
in proper range
Target compound (s)
99 -9# pure or of
known purity
See Subsection
3.6.10
Protect with suit-
able packing
material
Coiled and taped;
openings taped
Stored organized
in containers
Packed in rigid foam
in a rigid container
Packed in original
container, if pos-
sible, or rigid
container
See Section 3.3-3 of
this Handbook
See Section 3-1.3 of
this Handbook
Frequency and method
of measurement
Prior to field test
contact EPA or vendor
(see Subsection 3.6.8)
Prior to field test
contact manufacturer
or vendor
See Subsection 3.6.10
Prior to each shipment
As above
As above
As above
As above
As above
As above
Action if
requirements
are not met
Acquire required
audit gas(es)
Procure 99.9#
pure compound(s)
or compound(s) of
known purity
See Subsection
3.6.10
Repack
Coil and tape
Repack
As above
As above
See Section 3-3.3
of this Handbook
See Section 3-1.3
of this Handbook
(Continued)
-------
Section No. 3-16.3
Date June 30, 1988
Page 44
o
Table 3.2 (Continued)
Characteristic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Packaging Equip-
ment for Shipment
Glassware
Packed in original
shipping containers,
if available, or
suitable packing
material and marked
"Fragile"
Prior to each shipment
Repack
Tedlar bags
Packed in rigid sam-
pling containers, if
possible, or packed
individually in cor-
rugated boxes with
connectors secured
As above
As above
O
Sampling pumps
and dry gas meters
Mounted in a rigid
housing or packed in
rigid foam-lined
containers
As above
As above
Dilution interface
system
Built into a rigid
container suitable
for shipment
As above
Rebuild into
rigid container
or pack in suit-
able material
Gas chromatograph
system
Packed in original
shipping container,
secured properly in
van or trailer, or
mounted in a desig-
nated test vehicle
As above
Repack
Gas cylinders
Protective heads on,
secured in van or
trailer; transported
in compliance with
Federal, state, and
local regulations
As above
Repack; check
Federal, s tate,
and local regu-
lations concern-
ing transport of
compressed gases
Liquid organic
compounds
Top sealed and pack-
ed in original ship-
ping container
As above
Tape and repack
O
-------
Section No. 3.16.4
Date June 30, 1988
Page 1
4.0 ON-SITE MEASUREMENTS
On-site activities include transporting the equipment to the test site,
unpacking and assembling the sampling and/or analytical equipment, then conducting
the sampling and/or analysis for the predetermined organic compound(s). The qual-
ity assurance activities for the on-site measurements are summarized in Table 4.1
at the end of this section. Copies of all field data forms mentioned in this
section are in Subsection 3.16.12. The on-site measurements checklist. Figure 4.10
at the end of this section, provides the tester with a quick method for checking
requirements during sampling.
4.1 Transportation of Equipment to the Sampling Site
The most efficient means of transporting the equipment from ground level to the
sampling site (often above ground level) should be decided during the preliminary
survey or by prior correspondence. Care should be taken to prevent damage to the
equipment or injury to test personnel during the moving. A clean "laboratory" type
area free of excessive dust and organic compounds should be located and designated
for preparing the sampling systems and conducting sample recovery and analysis, if
applicable.
4.2 Preliminary Measurements and Setup
Method 18 strongly recommends that a preliminary survey and/or laboratory
evaluation be conducted prior to sampling and analysis. Unless adequate prior
knowledge of the source or information is available, the presurvey procedures
described in Subsection 3-0 on presampling operations should be followed to select
an acceptable sampling and analytical approach.
The accuracy of the sampling system(s) following handling and transportation to
the sampling site is determined using a cylinder gas audit. The integrity of the
system(s) is confirmed after setup by conducting the individual system check
described below for the applicable sampling method. Preliminary measurements will
always include determining the stack dimensions and the flue gas moisture. Other
measurements which may be made depending upon the requirements of the applicable
regulations and the source operations include a flow rate determination, velocity
check, and stack gas temperature range measurement.
One of the primary concerns for any organic sampling program must be safety.
The tester should always question the facility representative concerning general
plant safety requirements and safety in regard to sampling at the selected
sampling site. Every sampling and analysis protocol should address the safety
considerations involved in performing the protocol. Because there are numerous
safety considerations involved in organic sampling, it is beyond the scope of this
Handbook to discuss each one in detail. However, it cannot be over-emphasiged that
the tester must always be aware of the safety hazards.
4.3 Sampling
The following subsections discuss the procedures for each Method 18 sampling
technique. At this point, the tester has selected the proper sampling technique
and checked the selected sampling system. If this has not been accomplished, the
user should refer to Subsection 3-0 prior to conducting the field test.
-------
Section No. 3-16.1*
Date June 30, 1988
Page 2
Because of the complexity in sampling organic compounds from the variety of
potential source types, only the more common problems are addressed for each sam-
pling method. Recommended quality assurance/control checks and procedures are
provided to assess the suitability of the sampling technique for the samples to be
collected. Because of the relative compactness of the equipment and the low cost
of many of the sampling techniques, the tester may be able to utilize two different
sampling techniques at the same time with little additional effort. The samples
from the backup or secondary technique are not analyzed if the primary technique
proves satisfactory. For example, the tester might easily run an adsorption tube
system as a backup to an evacuated bag system. At some facilities, it may be
necessary to conduct two techniques simply to accurately measure all the organic
compounds of interest. The tester should always be aware that a change in process
operations such as raw materials, moisture content, operation mode, and temperature
can render a previously acceptable sampling technique unacceptable.
The specific sampling system descriptions are provided below.
4.3.1 Evacuated Container Sampling (Heated and Unheated) - In this procedure,
sample bags are filled by evacuating the rigid air-tight containers that hold them.
The suitability of the bags for sampling should have been confirmed by permeation
and retention checks using the specific organic compounds of interest during the
presurvey operations. The means of transporting the bags to the laboratory for
analysis within the specified time should also have been determined. Delays in
shipping and/or analysis can result in significant changes in concentration for
many compounds.
On-site sampling includes the following steps:
1. Conducting preliminary measurements and setup.
2. Preparation and setup of sampling system.
3. Preparation of the probe. ;
4. Connection of electrical service and leak check of sampling system.
5. Insertion of probe into duct and sealing of port.
6. Purging of sampling system.
7. Proportional sampling.
8. Recording data.
9. Recovering sample and transportion to laboratory.
Preliminary Measurements and Setup - The sampling site should be checked to
ensure that adequate electrical service is available. The stack dimensions are
measured and recorded on a data sheet similar to the ones shown in Figures 4.1, 4.2
and 4.3. The moisture content of the flue gas is used to correct the measured
concentrations to a dry basis. It is typically measured prior to sampling using
wet bulb/dry bulb thermometers or Method 4 (see Subsection 3^2); the determination
should be performed at a time when process operations are like they will be during
final sampling. If the process utilizes and emits ambient air, a sling psychro-
meter may be used to measure the moisture content of the ambient air in the area of
process air uptake. The moisture content value is also used to confirm that the
sampling approach selected is acceptable.
Prior to final sampling, the tester must determine if the final results are to
be presented on a concentration basis or a mass emission basis. If they will be
presented only on a concentration basis, only the concentrations of the specified
organics and the stack gas moisture content must be measured. If the mass emission
rate of any compound is to be presented, the flow rate of the stack gas using a
velocity traverse must also be determined. In this case, although not required by
Method 18, it is preferable that the sampling location be selected in accordance
/f-^.
f J
o
o
-------
Plant KubUr C-rU>n L*- Flowmeter calib.(Y) O. ?1 <> Dilution
City 7y^s/7tf., /VC-
Operator /4tf/?^ C^isve^
Date 5L/D.fc>/^
Run number' f?C.~l
Stack dia,
mm (in. ) (, /'M.
Sample box number A//A
Pitot tube (Cp)
f).f)4-
Static press Q mm (in.) H20
Sampling
time ,
min
O
/O
2-0
30
4-0
4-$
50
Total (fO
Clock
time,
24 h
W+l-
• '-
10:41
Velocity head
mm (in.) H20,
oSb
°0*&
°o%
"o.sz.
°^&
C/ •^f'\P
fj At
0.4k
Avg Q.SZ
Container type: bag X emissio
syringe diluent
canister
Container volume, 3<2
Container number
Average (aP) &.
Dilution
&&~~d>
^ jaa
Initial flowmeter setting
Average stack temp (?4
Barometric press
Flowmeter
setting
L/min (jPt*-/Eniii)
0- Ik!-
..-*.'£».
o.iw
0./6$-
0.141
0.14-1,
Av* 0.)&
liters emissio
Final le
"(in.) H20 Vacuum d
/.oz
?
?- mm titrT)
stack
U
JJ
iP 7
^ /
11
W
&
Avg (,*
probe
°C (°F)
MA
(
Avg
sy s tern : ( dynamic )
n flowsetting /V/4
flowsetting
system: (stat
n flowsetting
ak check ^
uring leak che
F) Sampling point locatio
Hg <%>9?&r- O/^tffavt-
tf/A
ic)
AS/A
m3/iain (cfm)
ck /O
cm (in.) H20
n
Temperature
sample line
°C (°F)
NlA
\
)
h
V,
1
Avg
readings
flowoeter box
°C (°F)
*/*
/
(
)
(
\
Avg
container
°C (°F)
#M
/
(
\
/
'
Avg
Figure 4.1. Field sampling data fora for container sampling.
nj o w
CD p CD
TO rt O
(D (D ft
LO
vD
O3
-------
Plant l\ubk-&f^ Cb\JL*f^_ Co-
City rfrsiH'c . tfC-
Operator /fan* Cl i>W6^
Date / /&4, Ibfr
Run number'
Stack dia.
/?d- / —*£d-f2--
nm (in.) ^ />,.
Barometric press T^T na (Iff.
Initial probe setting °C
"T Hg Sample loop volun
(°F) Sample loop temp
e i.O ml
-3./C? tO^(°F)
Sampling rate £. /5~2_ L/nin (cfm) Column temperature:
Sampling point local
C&fj^&r of &
ion
initial Cotisfant/
/CO °C/min
^^- program rate >A^ / °C/min
Dilution systen: final — / /OO °C/min
source flowrate 0.152. L/nin (cfn) Carrier gas flow
Meter box number /V/4
Stack temp
/gc>
?/<:
Vacuum @ check IO jrr"{in.) H00
ettings
diluent
nl/min
\
/
y
(
)
(
1
stack
ojg^jopj
/go
'/??
161
'm
/&D
%L
Temperat
probe
/80
l&b
/?0
izt
'ff*
/«
160
ure readings
sample line
ieo
I6o
'tel
^
/»/
'il'i
X
injection port
;
IT/
H
2ii
£-10
5if
3HO
Figure 4.2. Field sampling data form for direct interface sampling.
o
o
t) O C/5
{B S3 (D
m cr o
(0 (0 rt
H-
-p- c_i O
C 3
ro 2:
o
u.
CX>-
00-^
O
-------
Plant
City
Operator
Date /
Cklsn Co. Flowmeter calib. (Y) //2
Run number
Stack dia, mm (in.) (D //}.
Adsorption tube type:
charcoal tube X
silica gel
other
Dilution system: (dynamic)
emission flowsetting /V/4
diluent flowsetting
Meter box number
Pitot tube (Cp)
Static press ()
Adsorption tube number
Average (aP) Q.4- na (in.) H20
Initial flowmeter setting A/2.
Average stack temp ~7O
mm (in.) 1^0 Barometric press .2-9.65" ma (in.) Hg
Dilution system: (static)
emission flowsetting
Final leak check 0 m3/iain (cfin)
Vacuum during leak check /O
ma (in.) H.,0
Sampling point location '
Sampling
time,
min
O
/o
^
3o
45-
(sO
Total (fQ
Clock
time,
24 h
193*
,**
Velocity head
Sim-fin.) H20,
04
04
0.4*
04
04
0.4
04
04
0.4-
04
04
04
Avg Q q.
Flowmeter
setting
L/min (£ti/oln)
1.12-
1,/z
I.IZ
/,'/l
A/2
A/z
A/Z
1.12-
/./L
A/Z
1.1 1
Avg /, / Z_
stack
^6
tt
JO
*i
%.
7$
7-2.
Avg7/
Temperature r
probe, line
X
6?
(,1
70
^
?2
73
Avg T-/
•eadings
adsorp . tube
>C* (°F)
46
4?
%
fo
SI
si
Avg 5£>
meter
• o
4 9
5
52?
50
57
Avg S£>
Vacuum
^sf(in.) Hg
11
11
11
11
H
Z
11
Avg 11
tl O W
P {D fl>
W ft O
(0 (D ft
Figure 4.3. Field sampling data form for adsorption tube sampling.
O
U) •
O
- CO
vo cr>
oo.
OO.P-
-------
Section No. 3.16.4
Date June 30, 1988 x—^
Page 6 f J
with Method 1. If this is impractical, it should be selected to minimize flow
disturbances. The number and locations of sampling points for the velocity traverse
are selected according to Method 1 (see Section 3.0.1 of this Handbook); the trav-
erse is conducted according to Method 2 (see Section 3-1 of this Handbook). Note:
The Method 18 sampling will be conducted at a single point.
Method 18 requires that samples be collected proportionally, meaning that the
sampling rate must be kept proportional to the stack gas velocity at the sampling
point during the sampling period. If the process has a steady state flow (con-
stant) , then the flow rate does not have to be varied during sampling. The major-
ity of sources of organic emissions are of this type because they use constant rate
fans. If the tester can confirm from the facility that the source of interest has a
steady state flow (e.g., it uses a constant rate fan), then sampling can be conduc-
ted at a constant rate and no concurrent velocity measurements need to be made.
If it is not known whether the process is steady state or if it is not steady
state, then velocity measurements (the velocity head) must be made at the point to
be sampled. This can be done during the preliminary survey or before final
sampling, but should be done when the process operations are like they will be
during the final sampling. The average velocity head (pitot reading) and range of
fluctuation is determined and then utilized to establish the proper flow rate
settings during sampling. If it is found that the process is not steady state,
then the velocity head must be monitored during sampling to maintain a constant
proportion between the sample flow rate and the flow rate in the duct.
Select a total sampling time greater than or equal to the minimum total samp-
ling time specified in the applicable emission standard. The number of minutes
between readings while sampling should be an integer. It is desirable for the time
between readings to be such that the flow rate does not change more than 20% during
this period.
If it was determined from the literature or the preliminary survey laboratory
work that the sampling system must be heated during sample collection and analysis,
the average stack temperature is used as the reference temperature for the initial
heating of the system and should be determined. Then, the stack temperature at the
sampling point is measured and recorded during sampling to adjust the heating
system just above the stack temperature or the dew point. In addition, the use of
a heated sampling system typically requires that the analysis be conducted on-site
since it is not practical to maintain the sample bag at elevated temperatures for
long periods of time.
Sampling System Preparation - Prepare the probe and sampling train in the
laboratory area (see Figure 4.4). First, place a loosely packed filter of glass
wool in the end of the probe. Attach a sample bag that has been previously leak
checked to the sample container lid. Seal the inlet to the probe and the sample
container lid to the container body. Transport the container and probe to the
sampling site.
Proportional Sampling - Sampling must be conducted at a rate in constant
proportion to the stack gas flow at the sampling point. Thus, for a steady state
operation, the sampling flow rate is not varied during the run. For a non-steady
state process, the sampling flow rate is varied in proportion to the changing
velocity. The velocity is monitored by measuring the velocity head (AP) which is
linearly related to the square of the velocity. A recommended method for deter-
mining proportional sampling rates is as follows:
1. Conduct a single point velocity check as previously specified, and determine
the average velocity head (APa ) to be sampled.
o
-------
VENT
STACK
WALL
FILTER
(GLASS WOOL)
D
TEFLON
SAMPLE LINE
f
REVERSE
(3") TYPE
PITOT TUBE
PROBE
VACUUM LINE
MALE QUICK
PONNECTORS.
NEEDLE
VE
BALL
•f CHECK
FLOWMETER!
PUMP
CHARCOAL
TUBE
PITOT MANOMETER
NO CHECK
RIGID LEAKPROOF CONTAINER
Figure 4.4. Integrated bag sampling system.
T3 O W
jn p 0>
OT rt O
(D (0 rt
H-
-J «-« O
(0 "Z
O
t-o .
O
vO O\
00'
CD 4=-
-------
Section No. 3.16.4
Date June 30, 1988
Page 8
2. The average sampling flow rate for the test is determined prior to the start
of the run. Typically, the average sampling flow rate is about 0.5 L/min
which will yield approximately 30 liters of sample. The flow rate chosen in
the laboratory should fill the bag to about three fourths of its capacity
during the sample run. The average flow rate chosen is then assigned to the
average velocity head measured.
3. The flow rate to be used during sampling when the velocity head varies from
the average is calculated using the following equation.
Equation 4-1
Q. =
o
where
Q^ = Average sampling rate, L/min (ft3/min),
Qg = Calculated sampling rate, L/min (ft3/min),
AP = Actual velocity head, mm (in.) H20, and
APavg = Average velocity head, mm (in.) H20.
4. Determine the rotameter setting for the sampling rate (Qs) from the rota-
meter calibration curve, and adjust the rotameter accordingly.
Using this procedure will provide for the correct sampling rate and the proper
filling of the sample bag. Follow the procedure below to obtain an integrated
sample.
1. If a heating system is required, turn on the heating system and set at
average stack temperature determined from the pretest measurements.
2. Leak check the sampling train just prior to sampling by connecting a U-
tube, inclined manometer, or equivalent at the probe inlet and pulling a
vacuum of >^ 10 in. H20. Close the needle valve and then turn the pump off.
The vacuum should remain stable for at least 30 seconds. If a leak is
found, repair before proceeding; if not, slowly release the vacuum gauge.
This leak check is optional.
3- If the system is being heated, wait for it to come to the proper tempe-
rature. Place the probe in the stack at the sampling point: centroid of the
stack or no closer to the walls than 1 meter. Seal the sampling port to
prevent dilution of the stack gas by inleakage of ambient air.
4. Disconnect the flexible bag. Purge the system by turning on the pump and
drawing at least 5 times the sampling system volume through the train, or
purge for 10 minutes, whichever is greater.
5. Adjust the flow rate to the proper setting based on the velocity pressure
(during the purging, for non-steady state processes).
6. Connect the flexible bag to the sampling train (the connections should
ensure a leakfree system), and begin sampling. The rate must remain propor-
tional to the stack gas velocity for the total sampling time specified by the
standard of performance for the industry being sampled.
7. Record all data required (5 minute intervals, miniumum) on the field samp-
ling data form (see Figure 4.1). The flow rate and sample train heating
system should be adjusted after every pitot and temperature reading to the —
correct level. f ^
8. Disconnect and seal the flexible bag upon completion of sampling. Take care \ J
not to dilute the contents with ambient air.
9- Label each bag clearly and uniquely to identify it with its corresponding
data form and/or run. If the system is a heated system, the sample bag must
-------
Section No. 3.16.4
Date June 30, 1988
Page 9
be maintained at the stack temperature through sample analysis.
Sample Recovery and Transport to Laboratory - Sample recovery should be
performed so as to prevent contamination of the bag sample and maintain sample
integrity. The bag should remain leakfree, protected from direct sunlight, main-
tained at a temperature that will prevent condensation of any of the gases, and
stored in a safe place to prevent damage or tampering prior to analysis. It is
recommended that bag samples be analyzed within two hours of sample collection,
however, many of the organic compounds are stable enough to allow a few days prior
to analysis. Upon completion of the testing and sample recovery, all the data
forms should be checked for completeness and the sample bags reexamined for proper
identification.
Common Problems - The most common problems encountered with bag sampling
techniques are (1) adsorption of the gases on the bag, (2) permeation of the gases
through the bag, (3) reaction of gases in the bag, (4) condensation of the gases or
water vapor in the bag, and (5) leaks developing in the bag during testing, trans-
port, and/or analysis. As described previously in Subsection 3«0, the bags must be
checked for stability and retention of the compound in the bag. If the compound's
concentration significantly diminishes between the time the sample run is completed
and the time of analysis, then the bag technique will have to be modified or rejec-
ted. One modification that can be used to reduce both retention and/or condensa-
tion is addition of a heating system. Heating is generally applied during sample
collection and maintained through analysis. However, heating may increase the
permeation rate. Another option is the use of heat lamps applied to the sample
bags after sample collection and during sample analysis. Two other techniques
that have been used to prevent condensation are (1) addition of a knockout trap to
remove water vapor and heavy organics from the sample stream, and (2) use of
sorbents such as Tenax to remove the high boiling point organics. The tester must
demonstrate that the organic coiapound(s) of interest are not removed.
Alternatively, sample and/or water vapor condensation may be reduced by the use of
the prefilled« bag technique. The prefilling of the bag lowers the concentration
of the organic and/or water vapor, thereby eliminating condensation.
If the gases are reacting in the bag, then the bag material can be changed, the
time between sample collection and analysis reduced, or a different technique used
such as direct interface sampling. Methods to reduce bag leak problems are proper
construction of the sample bags, conducting additional runs, using a backup sample
collection technique such as an another bag sampling system or an adsorption tube
sampling system, and care with handling the sampling bags. Also, steel canisters
can be used in place of the bags. If the organic compounds are stable with time,
the use of steel canisters may better ensure the safety of the sample especially if
the samples must be air freighted to the laboratory for analysis.
4.3'2 Direct Pump Sampling - Direct pump sampling is conducted in a manner similar
to evacuated container sampling, with the exception that the needle valve and the
pump are located between the probe and sample bag and the sample exposed surfaces
of both must be constructed of stainless steel, Teflon or other material not affec-
ted by the stack gas (see Figure 4.5). Due to the additional likelihood that
sample may be lost in the needle valve and pump, it is recommended that the probe,
sample line, needle valve, and pump be heated. If it has or can be shown that this
not a concern, then the heating may be eliminated. All precautions, procedures,
data forms and criteria from Subsection 4.3-1 above can be applied. Ensure that
the system has been adequately purged before attaching the bag and sampling.
-------
Stainless
Needle Vafvo
Filter
(Glass Woo!)
Reverse
(3") Type
Pitot Tube
Rotamotar
Teflon-Lined
Diaphragm
Pump
O
Protective Container
Figure 4.5- Direct pump sampling system.
o
Vacuum
Lino
•T3 O W
o so ro
Oq rt O
fl> CO ct
H-
M e-< O
O C 3
ro 2:
o
co .
o
- OJ
vo cr\
oo •
o
-------
Section No. 3.16.4
Date June 30, 1988
Page 11
4.3-3 Explosion Risk Area Bag Sampling - Explosion risk area bag sampling is also
similar to evacuated container sampling. The major difference is that no electri-
cal components can be used in the explosion risk area. As previously mentioned in
Subsection 1.0, the first option of the tester is to locate the electrical equip-
ment (e.g., the pump) outside the explosion risk area and run a long flexible line
to the container. If that option is not possible, an evacuated steel container may
be used as shown in Figure 4.6. This option may involve a potential spark hazard
and must be checked though the plant safety officer. No electrical heating of the
system will likely be allowed. If an evacuated steel container is used, the leak
check can be conducted outside the explosion risk area and the probe can be purged
with a hand squeeze pump. The tester may wish to consider an alternative method of
sampling such as adsorption tubes and an" intrinsically safe personnel sampling pump
or the syringe method. The primary concern must be safety in an explosion risk area
and all operations must be outlined in writing and cleared through the Plant Safety
Officer. The same criteria as described above for suitability of the bag will
apply and must be met.
4.3.4 Prefilled Bag Sampling - The prefilled bag sampling technique is similar to
the heated direct pump sampling method. The major difference is that the sample
bag is prefilled with a known volume of nitrogen, hydrocarbon-free air, or cleaned,
dried ambient air prior to sampling and the volume of gas sampled must be accu-
rately determined (see Figure 4.5). When using a flowmeter or metering pump, the
maximum dilution that should be attempted is 10 to 1. Alternatively, a heated, gas
tight syringe may be used to collect the gas at the source and inject it into the
sample bag. A heated, gas tight syringe can be used for dilutions of 5 to 1 when
the dilution is performed in the syringe and 50 to 1 when performed in the bag.
The use of a heated, gas tight syringe should follow the procedures shown below in
Subsection 4.3«5. Both techniques should be verified•„in the laboratory using
higher concentrations of calibration gases and must be within 10% of the calculated
value. The technique is verified in the field by diluting the audit gases in the
same manner as the stack gases (see Subsection 8.0 for auditing procedures).
Following are the recommended steps to conduct prefilled bag sampling:
1. The sampling should be conducted proportionally as described above in Sub-
section 4.3.1. Calculation of the average sampling rate vs. the average P
will be the same with the exception that the volume of the prefilled inert
gas must be taken into account.
2. The suitability of the prefilled bag sampling technique should have been
checked in the laboratory. This would include calculating the dilution
factor required to obtain ;an acceptable sample concentration.
' 3« In the laboratory area, fill the sample bag (previously leak checked) with
the calculated volume of inert gas. Because of the potential for leaks,
bags should be filled the same day they are used. The inert gas volume
must be determined with a calibrated dry gas meter or mass flowmeter. The
bag should be sealed and taken to the sampling site.
4. At the sampling site, the sampling system is leak checked without the
sampling bag attached. Turn on the heating system and heat the system to
the stack temperature. Connect a U-tube H20 manometer or equivalent to the
inlet of the probe. After the system comes to the desired temperature, turn
on the pump and pull a vacuum of about 10 in. of H20. Turn off the needle
valve and shut off the pump. If there is no noticeable leak within 30
seconds, then the system is leak free.
5. Place the probe in the stack at the sampling point (centroid or no less
/i.
-------
PVC Tubing-
Directional
Nead!o
Vafvo
Quick Disconnectors
Probo
5'Triton
Tubing
o
Figure ^.6. Explosion risk area sampling system option using an
evacuated steel container.
co -c-
o
-------
6.
7-
8.
9.
Section No. 3.16.4
Date June 30, 1988
Page 13
than 1 meter from the wall) and seal the port so there will be no inleakage
of ambient air. Turn on the pump and purge the system for 10 minutes.
During the time that the system is purging, determine and set the proper
flow rate based on the Ap.
Turn off the pump and attach the sample bag. Compare the heating system
The sampling will be conducted proportionally. The stack temperature and
heating system temperature should be monitored and recorded. Record 'the
data on the sampling data form (Figure 4.1).
At the conclusion of the run, turn off the pump and remove the probe from
the duct. Remove the bag and seal it.
Conduct a final leak check. The system should pass the leak check; if it
does not pass, repeat the run.
4.3-5 Heated Syringe Sampling - The heated syringe technique can be used with the
prior approval of the Administrator. This technique should only be used when other
techniques are impractical. The heated syringe technique requires on-site analysis
with three syringes collected and analyzed for each run. The requirements for the
use of the syringes are the same as for the bag with regard to the reaction of the
gases with time and the retention of the gases in the syringe.
Following are the procedures recommended for the syringe sampling technique:
1. If heating is required, then the syringe must be encased in material that
has a high density to maintain the proper temperature. Alternatively, an
external heating system can be used that keeps the syringe at the proper
temperature until just before use and to which the syringe can be immedi-
ately returned.
2. The access port should be extremely small to prevent inleakage of ambient
air. The port may be covered with Teflon or other nonreactive material
that will allow the syringe to penetrate the material for sampling.
3. For the direct injection method (no dilution), place the syringe needle
into the stack and fill and discharge the full volume that will be sampled
three times. Then, draw the emission sample into the syringe, immediately
seal the syringe and return to the heating system, if applicable. The
second and third syringes are sampled at equal time intervals spanning the
required sample (run) time. The syringe samples must not be taken one
immediately after another.
4. For the diluted syringe method, the inert gas is introduced into the
syringe three times and discharged. Following this, the proper volume of
inert gas is pulled into the syringe. The syringe is then placed into the
duct and the proper volume of stack gas is added. Immediately remove the
syringe needle from the duct, seal the syringe, and return to the heating
system, if applicable.
5. For the bag diluted syringe method, the bag should be prefilled with the
proper volume of inert gas. The sampling is conducted as described above
and the sample injected into the bag through a septum.
6. Record the data on a field sampling data form (can adapt Figure 4.1).
7. Since the method requires a proportional sample to be collected, the
velocity head (AP) should be recorded at about the same time that each
sample is collected. The concentrations can then be mathematically
corrected to provide an integrated value. If the process is a constant
source operation (less than 10% change in flow over the sampling period),
it is not necessary to correct the measured values.
-------
Section No. 3.16.4
Date June 30, 1988
Page 14
4.3.6 Direct Interface Sampling - The direct interface procedure can be used
provided that the moisture content of the stack gas does not interfere with the
analysis procedure, the physical requirements of the equipment can be met at the
site, and the source gas concentration is low enough that detector saturation is
not a problem. Adhere to all safety requirements when using this method. Because
of the amount of time the GC takes to resolve the organic compounds prior to their
analysis, the GC can only typically make three analyses in a one-hour period.
Therefore, the number of injections in the direct interface method is greatly lim-
ited by the resolution time. At least three injections must be conducted per
sample run.
Following are the procedures recommended for extracting a sample from the
stack, transporting the sample through a heated sample line, and introducing it to
the heated sample loop and the GC. The analysis of the sample is described in
Subsection 5-0.
1. Assemble the system as shown Figure 4.7, making all connections tight.
2. Turn on the sampling system heaters. Set the heaters to maintain the stack
temperature as indicated by the stack thermocouple. If this temperature is
above the safe operating temperature of the Teflon components, adjust the
heating system to maintain a temperature adequate to prevent condensation of
water and organic compounds.
3. Turn on the sampling pumps and set the flow rate at the proper setting.
Typically 1 L/min is used.
4. After the system reaches the same temperature as the stack, connect a U-
tube H20 manometer or eqivalent to the inlet of the probe. Pull a vacuum
of about 10 in. of H20, and shut off the needle valve and then the pump.
The vacuum should remain stable for 30 seconds. If the system leaks,
repair and then recheck the system.
5. Calibrate the system as described in Subsection 5-0. Repeat until
duplicate analyses are within 5% of their mean value (Subsection 5«0).
6. Conduct the analyses of the two audit samples as described in Subsection
8.1. The results must agree within 10% of the true value (or greater, if
specified on the cylinder). If the results do not agree, repair the system
and repeat the analyses until agreement is met or until approval is given by
the representative of the Administrator.
7. After the audit has been successfully completed, place the inlet of the
probe at the centroid of the duct, or at a point no closer to the walls
than 1 meter, and draw stack gas into the probe, heated line, and sample
loop. Purge the system for a least 10 minutes.
8. Record the field sampling data on a form such as Figure 4.2.
9. Conduct the analysis of the sample as described in Subsection 5«0. Record
the data on the applicable data form (Figure 5-1, Subsection 5.0). Ensure
that the probe and sample lines are maintained at 0°C to 3°C above the
stack temperature (or a temperature which prevents condensation).
10. Conduct the posttest calibration as described in Subsection 5.0.
4.3.7 Dilution Interface Sampling - Source samples that contain a high concentra-
tion of organic materials may require dilution prior to analysis to prevent
saturating the GC detector. The apparatus required for this direct interface
procedure is basically the same as described above, except a dilution system is
added between the heated sample line and the gas sampling valve. The apparatus is
arranged so that either a 10:1 or 100:1 dilution of the source gas can be directed
to the chromatograph. The description of the apparatus is presented in Subsection
/"""N
f )
^—'
o
-------
MANOMETER
TC TC READOUT
READOUT OR CONTROLLER
GLASS
WOOL
1/2-in.
TUBING
1/4-in. SS
TUBING
STACK WALL
f—"*•» "~^ "VS"*™**.,.
mmm
-™V - >-ji___j—?
INSULATION
EMPERATUR
CONTROLLER
°| NEEDLE
VALVE
HEATED
[EFLON LINE
HEATED GAS
SAMPLING VALVE
SNGC
AUDIT
SAMPLE
EM
CHARCOAL
ADSORBER
FLOWMETER
PUMP
TO GC INSTRUMENT
CARRIER IN
Figure k.J. Direct interface sampling system.
TJ a en
o p a>
(ft ft O
n> a>
-------
Section No. 3-16.4
Date June 30, 1988
Page 16
1.1.9 and the pretest calibration of the apparatus is presented in Subsection
2.2.1.
Following are the procedures recommended for extracting a sample from the
stack, diluting the gas to the proper level, transporting the sample through a
heated sample line, and introducing it to the heated sample loop and the GC. The
analysis of the sample is described in Subsection 5.0.
1. Assemble the apparatus by connecting the heated box, as shown in Figure
4.8, between the heated sample line from the probe and the gas sampling
valve on the ehromatograph. . Vent the source gas from the gas sampling
valve directly to the charcoal filter, eliminating the pump and rotameter.
2. Measure the stack temperature, and adjust all heating units to a temper-
ature 0°C to 3°C above this temperature. If the temperature is above the
safe operating temperature of the Teflon components, adjust the heating to
maintain a temperature high enough to prevent condensation of water and
organic compounds.
3. After the heaters have come to the proper temperature, connect a U-tube H20
manometer or eqivalent to the inlet of the probe. Turn on the pump and pull
a vacuum of about 10 in. of H20. Shut off the needle valve and then turn
off the pump. The vacuum reading should remain stable for 30 seconds. If a
leak is present, repair and then recheck the system.
4. Verify operation of the dilution system by introducing a calibration gas at
the inlet of the probe. The diluted calibration gas should be within 10% of
the calculated value. If the results for the diluted calibration gas are
not within 10% of the expected values, determine whether the GC and/or the
dilution system is in error. If the analyses are not within acceptable
limits because of the dilution system, correct it to provide the proper
dilution factors. Make this correction by diluting a high concentration
standard gas mixture to adjust the dilution ratio as required.
5. Verify the GC operation using a low concentration standard by diverting the
gas into the sample loop and bypassing the dilution system as described in
Subsection 5-l« If these analyses are not within acceptable limits, correct
the GC by recalibration, etc.
6. Conduct the analyses of the two audit samples as described in Subsection
8.1 using either the dilution system or directly connect the gas sampling
valve as required. The results must agree within 10# of the true value or
greater value if specified on the cylinder. If the results do not agree,
repair the system and repeat the analyses until agreement is met or until
approval is given by the representative of the Administrator.
7. After the dilution system and GC operations are properly verified and the
audit successfully completed, place the probe at the centroid of the duct
or at a point no closer to the walls than 1 meter, and purge the sampling
system for at least 10 minutes at the proper flow rate. Conduct the analy-
sis of the sample as described in Subsection 5.0. Record the field and
analytical data on the applicable data forms (Figures 4.2 and 5-l)« Ensure
that the probe, dilution system, and sample lines are maintained at 0°C to
3°C above the stack temperature (or a temperature which prevents conden-
sation) .
8. Conduct the posttest calibration and verification of the dilution system as
described in Subsection 5.0.
If the dilution system is used for bag sampling, the procedures for verifying
operation of the dilution system will be the same as shown above. The diluted
calibration gas will be collected in a bag and then verified. Also the audit
samples will be collected in a bag and analyzed. Acceptable results must be
o
o
-------
Vent to Charcoal Adsorbers
Heated Line
from Probe
Quick
Connect
Quick
Connects to
Gas Sample
Valve
Source
Gas Pump
1.5L/Min
150 cc/Min
Pump
150 cc/Min
Pump
3-Way
Valves
in 100:1
Position
Check Valve
Quick Connects
for Calibration
Flowmeters
(On Outside
of Box)
Flow Rate of
1350 cc/Min
Heated Box at 120° C or Source Temperature
U
•ti o (/>
P p 0>
m rt o
(0 (0 rt
To Heated GC Sampling Valve
C
CD
OJ
o
Figure 4.8. Schematic diagram of the heated box required for
dilution interface sampling.
uo
v£> O\
co.
co-e-
-------
Section No. 3.16.4
Date June 30, 1988
Page 18 S~\
obtained for the audit samples prior to analysis of the field samples. >—'
4.3.8 Adsorption Tube Sampling - Adsorption tube sampling can be used for those
organics specified in the Method Highlights Section, Table B, and for other com-
pounds as specified in the National Institute of Occupational Safety and Health
(NIOSH) methods. The selection and use of adsorption tubes must be validated in
the laboratory as discussed in Subsections 3-3 and 3-4 or through the use of the
literature. This check will include selecting the proper adsorption material, and
then checking the capacity, breakthrough volume, adsorption efficiency, and desorp-
tion efficiency. The adsorption efficiency can be greatly affected by the presence
of water vapor and other organics in, and temperature of the stack gas. If sam-
pling is conducted for more than one organic compound, the adsorption and desorp-
tion efficiency checks must consider each. Because changes in process and control
equipment conditions can greatly affect all of the parameters stated above, it is
recommended as a standard operating procedure that more than one adsorption tube be
used. The first tube is analyzed as described in Subsection 5.0. If no problems
are found, then the second tube can be discarded. If problems with the first
tube's adsorption efficiency are discovered, then the primary section of the second
tube can still be analyzed and the results included with those of the primary
portion of the first tube.
Following are the recommended procedures for adsorption tube sampling:
1. The sampling system is assembled as shown in Figure 4.9- The adsorption
tube(s) must be maintained in a vertical direction for sampling. This is
done to prevent channeling of the gases along the side of a tube. It is
recommended that the sampling probe be eliminated when possible. If a /*""\
sample probe is used, it should be cleaned prior to its initial use with f J
the extraction solvent. Teflon tubing should be used for the probe and ^—
sample line.
2. Just prior to sampling, break off the ends of the adsorption tubes to
provide an opening at least one-half of the internal diameter. Audit
samples must be collected on the adsorption tubes during the test program
as described in Subsection 8.0. Since on-site analysis is typically not
conducted when using adsorption tubes, it is recommended that two samples
be collected from each of the two audit cylinders. This allows the tester
a second chance to obtain the proper value for each audit cylinder.
3. Prior to sampling and the collection of the audit samples, the sampling
system must be leak checked by connecting a U-tube H20 manometer or
equivalent to the inlet of the sample probe or adsorption tube. Turn the
pump on and pull a vacuum of about 10 in. of H20. Shut off the needle
valve and then turn off the pump. The vacuum must remain stable for 30
seconds. If a leak is present, repair and recheck the system.
4. If the flow rate in the duct varies by more than 10% during the sampling
period, the sample should be collected proportionally. The proportional
sampling procedures will be the same as described for the bag sampling. The
only difference is that instead of using the volume of the bag as the
limiting factor to determine the average sampling rate, the breakthrough
volume is the limiting factor. If the source is a constant rate source
(less than a 1Q% change in flow rate for the sampling period), the samples
can be collected at a constant rate.
5. Prepare the field blank just prior to sampling. The field blank will be
handled in be same manner as the field samples and should be from the same /"""N
lot as the other adsorption tubes. I }
6. The flow rate meter must have been calibrated in the laboratory prior to
-------
Probe
Supplemental
Adsorption
Tube
(as required)
Rotameter
Soap Bubble
Rowmeter
(for calibration)
A*
Sonic
Orifice
"o a en
CD fa n>
Oft} rt O
fl> CD ct
H-
M e_, o
vo C 3
Figure 4.9. Adsorption tube sampling system.
O
- CO
vo crv
co •
004=-
-------
Section No. 3• 16.4
Date June 30, 1988
Page 20
the field trip as described in Subsection 2.1. The volume of sample coll-
ected must be accurately known for adsorption tube sampling.
7. The sample run time must be equal to or greater than that specified by the
applicable regulation. During each sample run, the data should be recorded
on the sample data form {Figure 4.3 or equivalent).
8. At the conclusion of each run, conduct another leak check as described
above. If the system does not pass the leak check, the run should be
rejected, the leak located and repaired, and another run conducted.
9. After completing a successful leak check, remove the adsorption tube from
the holder and seal both ends with plastic caps. The tubes should be
packed lightly with padding to minimize the chance of breakage. If the
samples are to be held for an extended period of time, they should be kept
cool to reduce the amount of migration of the organic from the primary
section to the secondary section. Note: Pack the tubes separately from
bulk samples to avoid possible contamination.
10. It is recommended, that at the conclusion of the test, the sample probe (if
used) be rinsed into a 20-ml glass scintillation vial with about 5 to 10 ml
of the desorption solvent. This sample will be analyzed as a check on the
loss of the organic in the probe during sampling. If more than 10# of the
total sample collected in the adsorption tubes is present in the probe, the
samples should be rejected or the sample catch adjusted to account for the
loss. Alternatively, the probe can be rinsed after each run and the rinse
added to the desorption solvent prior to analysis.
11. At the conclusion of the test program, check all samples to ensure that
they are uniquely identified and check all data sheets to ensure that all
data has been recorded.
G
O
-------
Section No. 3.16.4
Date June 30, 1988
Page 21
WATER VAPOR CONTENT
Method 4
Reference Method conducted in proper manner (Handbook Section 3»3. Method
4, Figure 4.1)
Wet Bulb/Dry Bulb
Temperature readings taken when stabilized
WB Temp °C (°F) DB Temp °C (°F)
DIRECT OR DILUTION BAG SAMPLING
Apparatus
Pitot tube: Type S Other , Properly attached
Pressure gauge: Manometer Other , Sensitivity
Probe liner: Borosilicate Stainless steel .. . Teflon
Clean , Probe heater (if applicable) on Glass wool filter
(if applicable) in place Stainless steel or Teflon unions used
to connect to sample line
Sample line: Teflon , Cleaned , Heated (if applicable)
Bag: Tedlar Other , Blank checked , Leak checked
Reactivity check , Retention check
Flowmeter: Proper range , Heated (if applicable) , Calibrated
Pump: Teflon coated diaphram , Positive displacement pump ,
Evacuated canister , Personnel pump
Heated box with temperature control system: Maintained at proper temperature
Charcoal adsorption tube to adsorb organic vapors: Sufficent capacity
Dilution equipment: N2 gas , Hydrocarbon-free air , Cleaned and
dried ambient air , Dry gas meter
Barometer: Mercury , Aneroid , Other
Stack and ambient temperature: Thermometer , Thermocouple ,
Calibrated
Procedures
Recent calibration (if applicable): Pitot tube , Flowmeter ,
Positive displacement pump* , Dry gas meter* , Thermometer
Thermocouple , Barometer
Sampling technique: Indirect bag , Direct bag , Explosion risk bag
Dilution bag , Heated syringe , Adsorption tube ,
Proportional rate , Constant rate , Direct interface
Dilution interface
*Most significant items/parameters to be checked.
Figure 4.10. On-site measurements checklist.
-------
Section No. 3.16.4
Date June 30, 1988
Page 22 S~\
Figure 4.10 (Continued)
Filter end of probe (if applicable) and pitot tube placed at centroid of duct (or
no closer than 1 meter to stack wall) and sample purged through the probe and
sample lines* .
Vacuum line attached to sample bag and system evacuated until the flowmeter
indicates no flow (leakless)*
Heated box (if applicable) same temperature as duct*
Velocity pressure recorded and sample flow set
Proportional rate sampling maintained during run*
Stack temperature, barometric pressure, ambient temperature, velocity pressure
at regular intervals, sampling flow rate at regular intervals, and initial and
final sampling times recorded*
At conclusion of run, pump shut off, sample line and vacuum line disconnected
and valve on bag closed
Heated box (if applicable) maintained at same temperature as duct until analysis
conducted
No condensation visible in bag*
Sample bag and its container protected from the sunlight
Audit gases collected in bags using sampling system*
Explosive area bag sampling: (with following expections same as above)
Pump is replaced with an evacuated canister or sufficient additional line is added
between the sample bag container and the pump to remove the pump from the
explosive area
Audit gases collected in bags using sampling system*
Prefilled bag: Proportional rate Constant rate
Dilution factor determined to prevent condensation* ____
Proper amount of inert gas metered into bag through a properly calibrated dry gas
meter*
Filter end of probe (if applicable) and pitot tube placed at centroid of duct (or
no closer than 1 meter to stack wall) and sample purged through the heated probe,
heated sample line, and heated flowmeter or positive displacement pump*
Leak checked and partially filled bag attached to sample line
Stack temperature, barometric pressure, ambient temperature, velocity pressure at
regular intervals, sampling rate at regular intervals, and initial and final sam-
pling times recorded*
Probe, sample line, and properly calibrated flowmeter or positive displacement pump
maintained at the stack temperature*
Sampling conducted at the predetermined rate, proportionally or constant for entire
run*
No condensation visible in probe, sample lines, or bag*
At conclusion of run, pump shut off, sample line disconnected and valve on bag
closed
Sample bag and its container protected from sunlight
Audit gases collected in bags using dilution system*
o
Sample Recovery and Analysis
(As described in "Postsampling operations checklist," Figure 5-10)
*Most significant items/parameters to be checked.
O
-------
Section No. 3.16.4
Date June 30, 1988
Page 23
Figure 4.10 (Continued)
DIRECT AND DILUTION INTERFACE
Apparatus
Probe: Stainless steel , Glass , Teflon , Heated system (if
applicable) , Checked
Heated sample line: Checked*
Thermocouple readout devise for stack and sample line: Checked*
Heated gas sample valve: Checked*
Leakless Teflon-coated diaphram pump: Checked*
Flowmeter: Suitable range
Charcoal adsorber to adsorb organic vapors
Gas chromatograph and calibration standards (as shown in "Postsampling operations
checklist," Figure 5-10)*
For dilution interface sampling only:
Dilution pump: Positive displacement pump or calibrated flowmeter with Teflon-
coated diaphram pump checked*
Valves: Two three-way attached to dilution system
Flowmeters: Two to measure dilution gas, checked*
Heated box: Capable of maintaining 120°C and contains three pumps, three-way
valves, and connections, checked*
Diluent gas and regulators: N2 gas , Hydrocarbon-free air , Cleaned air _,
Checked
Procedures
All gas chromatograph procedures shown in "Postsampling operations checklist"
(Figure 5-10)
Recent calibration: Thermocouples , Flowmeter , Dilution system
(for dilution system only)*
Filter end of heated probe placed at centroid of duct (or no closer than 1 meter to
, stack wall), probe and sample line heat turned on and maintained at a temperature
of 0°C to 3°C above the source temperature while purging stack gas
Gas chromatograph calibrated while sample line purged*
After calibration, performance audit conducted and acceptable*
Sample line attached to GC and sample analyzed after thorough flushing*
With probe removed from stack for 5 min, ambient air or cleaned air analysis is
less than 5# of the emission results*
Probe placed back in duct and duplicate analysis of next calibration conducted
until acceptable agreement obtained*
All samples, calibration mixtures, and audits are analyzed at the same pressure
through the sample loop*
Sample Analysis
(As shown in "Postsampling operations checklist," Figure 5-10)
*Most significant items/parameters to be checked.
-------
Figure 4.10 (Continued)
Section No. 3.16.4
Date June 30, 1988
Page 24 ^^^
If a dilution system is used, check the following:
With the sample probe, sample line, and dilution box heating systems on, probe and
source thermocouple inserted into stack and all heating systems adjusted to a
temperature of 0°C to 3°C above the stack temperature
The dilution system's dilution factor is verified with a high concentration gas of
known concentration (within 10%)
The gas chromatograph operation verified by diverting a low concentration gas into
sample loop
The same dilution setting used throughout the run
The analysis criteria is the same shown as for the direct interface and in the
"Postsampling operations checklist," Figure 5-10
ADSORPTION TUBES
Apparatus
Probe: Stainless steel , Glass , Teflon , Heated
system and filter (if applicable)
Silica gel tube or extra adsorption tube used prior to adsorption tube when
moisture content is greater than 3 percent
Leakless sample pump calibrated with limiting (sonic) orifice or flowmeter
Rotameter to detect changes in flow
Adsorption tube: Charcoal (800/200 mg), Silica gel (1040/260 mg) S~*\
Stopwatch to accurately measure sample time f j
Procedures
Recent calibration of pump and flowmeter with bubble meter
Extreme care is taken to ensure that no sample is lost in the probe or sample line
prior to the adsorption tube
Pretest leak check is acceptable (no flow indicated on meter)
Total sample time, sample flow rate, barometric pressure, and ambient temperature
recorded ._
Total sample volume commensurate with expected concentration and recommended sample
loading factors
Silica gel tube or extra adsorption tube used prior to adsorption tube when
moisture content is greater than 3 percent
Posttest leak check and volume rate meter check is acceptable (no flow indicated on
meter, posttest calculated flow rate within 5 percent of pretest flow rate)
Sample Analysis
(As shown in the "Postsampling operations checklist," Figure
*Most significant items/parameters to be checked.
o
-------
Section No. 3
Date June 30,
Page 25
1988
Table 4.1. ACTIVITY MATRIX FOR ON-SITE MEASUREMENTS
Characteris tic
Preliminary de-
terminations
and measure-
ments
Acceptance limits
If final results on
concentration basis,
determine the mois-
ture content of stack
gas
If final results on
emission rate basis,
determine moisture
content and flow
rate of stack gas
If process has >10%
variation in APave,
sampling must be con-
ducted proportionally
If preliminary survey
or information showed
a heating system nec-
essary for sampling,
determine stack gas
temperature, Ts
Determine stack
dimensions
Select sampling time
minimum total
sampling time in •>
applicable emission
standard; number or
minutes between
readings should be an
integer
Frequency and method
of measurement
Once each field test;
use wet bulb/dry bulb
thermometer, Method 4,
or sling psychrometer
See above for moisture
content; for flow
rate, once each field
test using Method 1
location, if possible,
and Method 2 proce-
dures
Determine before
test by measuring
APave and range of
fluctuation; if re-
quired, use APave and
AP measured during
sampling to vary
sampling flow rate to
sample proportionally
• Prior to and during
sampling
Prior to sampling,
using tape measure
Prior to sampling
Action if
requirements
are not met
Complete
Complete
Complete or
repeat sampling
Complete or
repeat sampling
Complete
Complete
(Continued)
-------
Table 4.1 (Continued)
Section No. 3.16.4
Date June 30, 1988
Page 26
o
Characteristic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Evacuated con-
tainer sampling
1. Assemble system
using Fig. 4.4;
leakage
no
2. Minimum vacuum
of 10 in. of H20;
stable for 30 s
3. Heating system, if
used, between 0°
and 3°C above
4. Locate probe tip
at centroid of
stack or no closer
than 1 meter to
walls of stack
5. Purge probe and
sample system,
5 times system
volume or 10
minutes , which-
ever is greater
6. Sample propor-
ionally based on
APove and moni-
tored AP
Before sample collec-
tion, visually and
physically inspect
all connections
Before sample collec-
tion; use a
filled U-tube
manometer or equiva-
lent
Confirm prior to and
monitor during sam-
pling using tempera-
ture sensor(s)
Prior to sampling;
determine using stack
dimensions
Before sample collec-
tion; with bag
unattached, turn on
pump
Throughout sampling
Check for leaks,
repair system;
repeat check
Check system
for leaks;
check pump,
joints, and
valves for source
of leak; repair
and recheck
Adjust heating
system
Reposition
O
Repeat purge
Repeat test
Direct pump
sampling
(Continued)
Assemble system
using Fig. 4.5;
sample exposed
components of
Teflon, stain-
less steel, etc;
no leakage
Before sample col-
lection, visually
and physically in-
spect all equipment
and connections
Check for leaks,
repair system;
replace inappro-
priate components
O
-------
Table 4.1 (Continued)
Section No. 3.16.4
Date June 30, 1988
Page 27
Characteristic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Direct pump
sampling (cont)
2. Follow steps 2-6
for evacuated bag
sampling
As above
As above
Explosion risk
area bag
sampling
1. Assemble system,
Figure 4.6 is one
option; no elec-
trical compo-
nents in explo-
sion risk area;
no leakage
2. Leak check as
above outside
explosion risk
area
3« Purge probe with
a hand squeeze
pump changing
volume at least
5 times
4. Follow steps 4
and 6 for evac-
uated bag samp-
ling
5. Clear all oper-
ations in writ-
ing through
Plant Safety
Officer
As above
As above
As above
As above
As above
As above
As above
As above
Prior to working
in explosion risk
area
Complete
Prefilled bag
sampling
(Continued)
1. Assemble system
using Fig. 4.5;
need calibrated
flowraeter in-line
2. Calculate accept-
able dilution
factor
3. Leak check bag
As above for
evacuated bag
sampling
Prior to sampling
Prior to filling
As above for
evacuated bag
sampling
Complete
Repair or replace
-------
Table 4.1 (Continued)
Section No. 3.16.4
Date June 30, 1988
Page 28
o
Characteristic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Prefilled bag
sampling (cont)
4. Fill bag with
known volume of
inert gas
5. Leak check system
at stack temper-
ature , minimum
vacuum of 10 in.
of H20, stable
for 30 s
6. Follow steps 3~6
for evacuated bag
sampling
7. Determine volume
of gas sampled
accurately
Prior to sampling;
use calibrated dry gas
meter or mass flow-
meter
Before sample collec-
tion without bag
attached; use U-tube
H20-filled manometer
or equivalent
As above
During sample collec-
tion; use flowmeter
or metering pump (max.
dilution 10 to 1) or
heated syringe, (see
below (max. dilution
50 to 1)
Complete
Locate leak,
repair or
replace compo-
nents , and
recheck
As above
Complete
O
Heated syringe
sampling -
direct injec-
tion
1. Check syringes
for compound re-
tention and re-
action
2. Seal port to pre-
vent inleakage
of ambient air
3- Place needle in
stack at sample
point, pull and
discharge sample
volume three
times
See Subsection 1.0
Visually check
Prior to sampling
Complete
Reseal and re
check
Complete
(Continued)
O
-------
Table 4.1 (Continued)
Section No. 3.16.4
Date June 30, 1988
Page 29
Characteris tic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Heated syringe
sampling - dir-
ect injection
(cont)
Seal after pull-
ing sample vol-
ume, return to
heating system,
if necessary;
monitor heating
system tempera-
ture
Choose sample
volumes to sample
proportionally;
monitor AP, if
necessary (>10#
change in flow
over sampling
period)
Take second and
third syringe
samples at equal
time intervals
spanning the
required sampling
time in applica-
ble emission
standard
For each sample
collection; use
temperature sensor
Complete
During sample collec-
tion; use pitot tube
Repeat sampling
During field test
Repeat sampling
Heated syringe
sampling -
dilution
method
(Continued)
1. Follow same steps
as for heated
syringe - direct
injection, except
prefill bag (see
steps 2-4 in pre-
filled bag samp-
ling) and inject
gas in heated
syringe through
bag septum
As above
As above
-------
Section No. 3.16.4
Date June 30, 1988
Page 30
Table
(Continued)
o
Characteristic
Acceptance limits
'Frequency and method
of measurement
Action if
requirements
are not met
Direct interface
sampling
1. Assemble system
using Fig. 4.7;
no leakage
2. Heating system
between 0° and
3°C above T
3. Set flow rate at
1 L/min
4. Leak check system
at stack temper-
ature at minimum
vacuum of 10 in.
of H20; stable
for 30 s
5. Calibrate system;
duplicate analy-
ses within 5% of
their mean
6. Analyze audit
cylinders;
results within
10% of true value
7. Follow steps 4
and 5 of evacu-
ated bag
sampling
8. Analyze samples
and conduct
posttest
calibration
Before sample collec-
tion visually and
physically inspect
all connections
Confirm prior to and
monitor during
sampling using temper-
ature sensors
Prior to sampling
During sampling; use
a U-tube H20 mano-
meter or equivalent
See Subsection 5«0
See Subsection 8.0
As above
See Subsection 5-0
Check for leaks;
repair system;
repeat check
Adjust
Complete
Check system for
for leaks; repair
and recheck
Identify
problems; recal-
ibrate and check
Reject samples
and rerun test
As above
Complete
O
(Continued)
O
-------
Section No. 3.16.4
Date June 30, 1988
Page 31
Table 4.1 (Continued)
Characteristic
Dilution inter-
face sampling
Acceptance limits
1. Follow all steps
.. for direct inter-
face sampling
with addition
of steps below
2. Assemble with
dilution system
in line, see
Figure 4.8
3. If Teflon com-
ponents cannot
withstand stack
temperature,
heating system
should be set to
prevent conden-
sation
4. Verify dilution
system to within
10# of expected
value
Frequency and method
of measurement
As above
As above
Prior to and during
sampling
Prior to sampling; use
a calibration gas
Action if
requirements
are not met
As above
As above
Adjust
Pinpoint problen
to dilution
system or QC;
repair and
recheck; adjust
dilution, if
necessary
Adsorption tube
sampling
i. Assemble system
using Figure 4.9
2. Break off ends of
adsorption tubes;
maintain in ver-
tical position
for sampling
3. Follow step 4 for
direct interface
for leak check
Before sample coll-
ection, visually and
physically check all
connections
Just prior to samp-
ling; during sampling
As above
Check for leaks,
repair, and
recheck
Complete and
check
As above
(Continued)
-------
Table 4.1 (Continued)
Section No. 3.16.4
Date June 30, 1988
Page 32
o
Characteristic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Adsorption tube
sampling (cont)
4. Conduct sampling
proportionally,
if the flow rate
varies more than
10%
5. Determine samp-
ling time, >^ re-
quirement of
applicable
emission standard
6. Run field blanks
7- Perform posttest
leak check as
above
During sampling
Repeat test
Prior to sampling
Complete
Once each set of
samples
As above
Complete
Reject sample
rerun test
• o
Sample
recovery
(Continued)
1. If applicable,
remove samples
from sampling
system
2. Protect bag samp-
les from sunlight
and maintain at a
temperature which
will prevent con-
densation
3. Analyze bag sam-
ples within two
hours of sampling
4. For adsorption
tube samples,
perform at least
one probe rinse
with desorption
solvent to con-
firm that <10% of
sample is col-
lected in probe
Following sampling
Following sampling
Complete
Complete
Following sampling
Following sampling;
analyze sample with
GC
Complete
Adjust sample
values to
account for probe
catch
O
-------
Section No. 3.16.4
Date June 30, 1988
Page 33
Table 4.1 (Continued)
Characteristic
Sample
logistics
Acceptance Limits
Properly label
all bags, contain-
ers, tubes, etc.
Record all data on
forms in Figs. 4.1,
4.2, and 4.3 and
5.1
Frequency and method
of measurement
Visually check
each sample
As above
requirements
are not met
Complete the
labeling
Complete the
data records
-------
o
o
o
-------
Section No. 3.16.5
Date June 30, 1988
Page 1
5.0 POSTSAMPLING OPERATIONS
The postsampling operations for Method 18 include preparation of calibration
standards appropriate for the sampling technique used, determination of desorption
efficiency and collection efficiency for adsorption tubes (if used), adsorption
tube sample preparation, sample analysis, and determination of;acceptable resolu-
tion and precision. See Subsection 3-1.5 f°r postsampling operations related to
velocity determinations and Subsection 3-3'5 for postsampling operations related to
determination of the flue gas."pr duct moisture. Figure 5-10 at the end of this
section provides a checklist for monitoring the postsampling operations. Table 5'1
at the end of the section summarizes the quality assurance activities associated
with the postsampling operations.
5.1 Preparation of Calibration Standards
Calibration standards are to be prepared prior to sample analysis following
the procedures described in the following subsections. Refer to Table E in the
Method Highlights Section for recommendations on the procedures suitable for selec-
ted compounds. Note that there are two basic types of standards, gaseous or liq-
uid; the type prepared depends on the type of sample collected. Gaseous cali-
bration standards will be needed prior to the analysis of preliminary survey sam-
ples collected in glass flasks or bags, and final samples collected in bags or by
direct and dilution interface sampling. There are three techniques for preparing
gaseous standards, depending on availability and the chemical characteristics of
the standard compound(s); gas cylinder standards may also be used directly, if the
proper concentration ranges are available. Liquid calibration standards are re-
quired for the analysis of adsorption tube samples from the preliminary survey
and/or the final sampling, as well as to determine the desorption efficiency; there
are two techniques for preparing liquid calibration standards. The concentrations
of the calibration standards should bracket the expected concentrations of the
target compound(s) at the source being tested. Specific procedures for preparing
and analyzing each type of standard are described below.
For each target compound, a minimum of three different standard concentra-
tions are required to calibrate the GC. An exception to this requirement involves
developing relative response factors for each compound to be tested as compared to
a single organic compound. Once in the field, the GC is calibrated for all target
compounds using the single organic. The validity of this procedure must be first
be proven in the laboratory prior to the test. To save time, multiple component
standards can be prepared and analyzed provided the elution order of the components
is known.
It is recommended that the linearity of the calibration curve be checked
comparing the actual concentration of the calibration standards to the concentra-
tion of the standards calculated using the standard peak areas and the linear re-
gression equation. The recommended criteria for linearity ,is for the calculated
concentration for each standard be within 7% of the actual concentration.
After establishing the GC calibration curve, an analysis of the audit cyl-
inder is performed as described in Subsection 8.1. For an instrument drift check,
a second analysis of the calibration standards and generation of a second
calibration curve is required following sample analysis. The area values for the
first and second analyses of each standard must be within 5# of their average. If
this criterion cannot be met, then the sample values obtained using the first and
second calibration curves should be averaged. In addition, if reporting such
-------
Section No. 3-16.5
Date June 30, 1988
Page 2
average values for the samples is warranted, an additional analysis of the audit
cylinder should be performed. The average of the audit values obtained using the
two calibration curves should be reported.
5.1.1 Commercial Gas Cylinder Mixtures - Commercial gas cylinder mixtures can be
used provided that the cylinders have been certified by direct analysis and the
proper concentrations for the emission test can be obtained. Calibrate the GC
using gas cylinders by the following procedure:
1. Secure the three cylinders xnear the GC and remove their protective caps.
Attach an appropriate regulator that is equipped with a flow control valve
to the lowest concentration standard.
2. For preliminary survey sample analysis, establish the proper GC conditions
based on the referenced conditions in Table D in the Method Highlights
Section, previous experience, or possibly, if the plant being tested has a
laboratory, the laboratory personnel. For final sample analysis, establish
the optimum GC conditions determined during the preliminary survey sample
analysis.
3. Attach a quick connect or equivalent, compatible to the connection on the
Tedlar bag or the interface sample line, to the gas sampling valve on the
GC.
k. Connect a length of Teflon tubing to the flow control valve on the regula-
tor and connect the other end, using a compatible connector, to the gas
sample valve.
5- With the gas sampling valve in the load position and the flow control valve
open, open the valve on the cylinder and adjust the pressure regulator to
deliver a flow of 100cc/min through the sample loop, determined by a rota-
meter or other flow sensing device on the loop outlet.
6. Allow the sample loop to be flushed for 30 seconds, then turn off the flow
control valve.
7. Allow the sample loop to return to the same pressure that will be exper-
ienced during sample analysis, determined with a manometer or equivalent
connected to a tee on the outlet of the loop, and immediately switch the
valve to the inject position.
8. Note the time of the injection on the strip chart recorder and/or actuate
the electronic integrator. Also, record the standard concentration, detec-
tor attenuation factor, chart speed, sample loop temperature, column tem-
perature and identity, and the carrier gas type and flow rate on the data
form shown in Figure 5-l» It is also recommended that the same information
be recorded directly on the chromatogram. Record the operating parameters
for the particular detector being used.
9- After the analysis, determine the retention time of each standard component
and determine the peak area. Repeat the injection of the first cylinder gas
standard until the area counts from two consecutive injections are within 5
percent of their average.
10. Multiply the average area count of the consecutive injections by the atten-
uation factor to get the calibration area value for that standard concen-
tration. NOTE: Attenuation factors which affect the plot traced, but not
the area count returned by an electronic integrator should not be
multiplied by the average area count. Observe the effect of attenuation
changes made at the console of a specific electronic integrator to
determine the appropriate course of action.
o
o
o
-------
Section No. 3.16.5
Date June 30, 1988
Page 3
Analysis of Method 18 Field Samples
/&& Analyst: JT
Date:
Location:
Type of Calibration Standard:
Number of Standards: 3 Date Prepared:
Plant:
Sample Type:
Target Compound:
Prepared By: A-
OC Used: HP
Column Used: Ps>i
Calibration Data
First analysis/second analysis
Standard concentration (Cact)
Flow rate through loop (ml/min)
Liquid injection volume (tubes)
Injection time (24-hr clock)
Chart speed (cm/min)
Detector attenuation
Peak retention time (min)
Peak retention time range (min)
Peak area
Peak area x attenuation factor
Average peak area value (Y)
Percent deviation from average
Calculated concentration (Cstd)
% deviation from actual (#Dact)
Linear regression equation; slope
Sample Analysis Data
First analysis/second analysis
Sample identification
Interface dilution factor
Flow rate through loop (ml/min)
Liquid injection volume (tubes)
Injection time (24-hr clock)
Chart speed (cm/min)
Detector attenuation
Peak retention time
Peak retention time range
Peak area
Peak area x atten. factor (Aj/A2)
Average peak area value (Y)
% deviation from average (#Davg)
Calculated concentration (Cs)
(Y - b) At
m
Standard 1
/O.O
7,00! i.oc?
NA- 1 M
l4-'Oo
te> 1 /o
2-/ -Z-
2.03/2.62.
o.ol
2-I& / ZO&
43£/ 4-K*
4-1b
2.3
/0.2,
-3-0
(m): 4/.6P
Sample 1
££-/
A/A
2-<7O/ 2JOQ
V4 / AfA-
IS:Zt,l IS;W
It) I /O
2, 1 T-
1.61/Z.&3
O-°l
BOA-I t>5t>
/(tOQ/ 1 <*?<<,
KotZ.
J-.l
39.3
- Y
Y
Standard 2
30-0
Z-£>O 1 £,£>
A/ A- / A/A
/o 1 /o
Z- 1 2-
Z.£Z/ 2.6f-
O.OZ,
(,/+/ ffZ&
IZZB//i?(s
/Z.4-2.
/./
2.1.7-
-1,0
y-intercept
Sample 2
AC- 2
fi/A
2oo / ZOO
A/frl AJA-
Hr'.l&l J(*'-+3
/o 1 ID
2- / 2-
2.g//2.g3
O.toZ.
/OW,I /0?4-
jLlBZ/ ZJ4-6
-L\4o
0.4-
o
o
(b): _ o.BI
Sample 3
A £-3
Afft
2-00/2.0O
A/A-/ ASA
1 ?-:zt> I '?'•*<*>
/O / /O
z. 1 i-
2.86/2 -go
«
?ex]1i?-
/g/£j//534-
/8<2fc
0.4-
4-3.1-
_ fi
t d act
Figure 5.1. Data form for analysis of Method 18 field samples.
-------
Section No. 3.16.5
Date June 30, 1988
Page 4
11. Repeat the procedure for the other standard concentrations, with the high-
est concentration analyzed last.
12. Prepare a plot with the standard concentration (Ca) along the abscissa (x-
axis) versus the corresponding calibration area value along the ordinate
(y-axis). Perform a regression analysis to calculate the slope and the y-
intercept. Draw the least squares line on the plot.
5.1.2 Preparation and Analysts of Gaseous Standards from High Concentration Cylin-
ders - Gaseous standards can be prepared from high concentration cylinder gases by
dilution with hydrocarbon-free gas and collection of the diluted gas mixture in a
Tedlar bag (10 liters or larger). A single-stage dilution system is used for
dilutions up to about 20-fold. For greater dilutions, a two-stage dilution system
should be used. It is recommended that a check of the dilution system be made by
analyzing a low concentration cylinder standard that is in the range of one of the
standards prepared by dilution. Prepare the gaseous standards, by the dilution
technique, using the procedures that follow:
1. Assemble the single-stage dilution system, as shown in Figure 5-2 and/or
the two-stage dilution system, as shown in Figure 5-3, using rotameters
(flowmeters) calibrated following the procedures described in Subsection
2.1.3- (More precise dilutions may be possible if the dilution system
utilizes mass flow controllers and mass flowmeters in place of the
rotameters.)
2. Connect the primary flowmeters on the single-stage system to the calibra-
tion gas mixture and the diluent gas (hydrocarbon-free). On the two-stage
system, connect the secondary flowmeters to the two diluent gas cylinders.
3- Connect a leakfree evacuated Tedlar bag fitted with a quick connect or
equivalent, compatible to the connection on the actual sample bags or the
interface sample line, to the tee connector on the single-stage system or
following the second stage of the two-stage system.
4. Open the gas cylinders, adjust all the pressure regulators to the same
pressure, and adjust the gas flows to achieve the desired dilution ratio
using the flow control valves. On the two-stage system, adjust the needle
valve on the high concentration waste outlet so that 9Q% of the high con-
centration gas is wasted and 10# goes to the second stage. NOTE: Divert
high concentration waste to a fume hood or pass it through an appropriate
adsorption media to protect personnel from exposure to harmful
concentrations of organic vapors.
5- Take periodic readings of the pressure difference between the first and
second stages of the two-stage system, as indicated by a water manometer or
equivalent, to correct the flow reading from the first stage to the second
stage. If the flow rates of the two stages can be balanced so that the
pressures are equal, then no correction will be necessary.
6. Disconnect the Tedlar bag from the dilution system before the bag is total-
ly full, and turn off the gases. Label the bag to indicate the contents,
the time and data when it was prepared, the identity of the high concentra-
tion gas cylinder, and the dilution ratio(s) used.
7- Record the ambient temperature, the flow meter readings, the barometric
pressure, and the average first stage pressure on the data form shown in
Figure 5.4.
8. Calculate the concentration (Cs), in ppmv, of the component in the final
gas mixture by the following formulas for 'single-stage and two-stage
dilution.
o
o
-------
COMPONENT
GAS
CYLINDER
DILUENT
GAS
CYLINDER
"T" CONNECTOR
CALIBRATED ROTAMETERS
WITH FLOW CONTROL
VALVES
"H
TEDLAR BAG
O OT
P fl>
Ct O
-------
MANOMETER
HIGH
CONCENTRATION
WASTE
NEEDLE VALVES
ROTAMETERS
LOW
£»• CONCENTRATION
GAS
DILUENT AIR
PURE SUBSTANCE OR
PURE SUBSTANCE/N, MIXTURE
Figure 5.3. Two-stage calibration gas dilution system.
13 O to
O P CD
«-i O
og:
OJ .
O
• OO
OS-
OOVJI
O
O
O
-------
Section No. 3.16.5
Date June 30, 1988
Page 7
Preparation of Standards by Dilution of Gas Cylinder Standards
Date:
2/Z-t/
Preparer:
Purpose:
Cylinder Component :
Source:
_ _
Component Concentration (X) :
-------
Section No. 3.16.5
Date June 30, 1988
Page 8
o
For single-stage dilution:
(X x qc)
Cs = Equation 5-1
where
X = Mole or volume fraction of the organic in the calibration gas
that was diluted, ppmv,
qc = Flow rate of the calibration gas that was diluted, and
qd = Diluent gas flow rate.
For two-stage dilution:
1c2 corr = <*c 2 actual X ~ ~ Equation 5-2
where
^c2 corr = Corrected flow rate from the first stage to the second stage,
qc2 actual = Actual flow rate from the first stage to the second stage,
Pd = Average differential pressure between the first and second j^*\
stage, mm or in. H20, and f j
Pb = Barometric pressure, mm or in. H20. X_x
and
"c1 ^c 2 corr
Cs = X x x Equation 5-3
(qcl * qdl) (qc2 corr + qd2)
where
X = Mole or volume fraction of the organic in the calibration gas
that was diluted,
qcl = Flow rate of the calibration gas diluted in the first stage,
qdl = Flow rate of the diluent gas in the first stage, and
qd2 = Flow rate of the diluent gas in the second stage.
9. Prepare two more calibration standards from the high concentration cylin-
der gas sufficient to bracket the expected concentration in the source
samples.
10. Analyze the calibration standards following the procedures described in
Subsection 5«1«1 for commercial gas cylinder mixtures. Use a pump on the
outlet side of the sample loop to flush the standards through the loop at
100 cc/min for 30 seconds. Using a manometer connected to a tee on the
outlet of the sample loop, make certain that the sample loop pressure
during analysis of the calibration standards is equal to the loop pressure
during actual sample analysis.
11. Once the calibration curve is established, it is recommended, if available,
that an undiluted cylinder standard in the range of the standard curve be
analyzed to verify the dilution ratio. Analyze the cylinder and calculate
the sample area value by multiplying the peak area by the attenuation
O
-------
Section No. 3.16.5
Date June 30, 1988
Page 9
factor. Use the slope and the y-intercept derived from the linear regres-
sion equation and the sample area value to calculate the cylinder concen-
tration (Cs) by the following formula:
Y - b
C. =
Equation 5-4
where
Y = Sample peak area, area counts,
b = y-intercept of the calibration curve, area counts, and
S = Slope of the calibration curve, area counts/ppmv.
The calculated concentration of the undiluted cylinder standard, based on the
analysis and the calibration curve generated from the diluted standards, should be
within 10% of the true value of the undiluted cylinder. If this criteria cannot be
met, then the GC calibration should be checked, the diluted sample may be outside
the calibration range, or there is a problem with the dilution system used to
prepare the standards (e.g. the rotameters are out of calibration, etc.)- Identify
the problem and correct it, or use one of the other approaches for preparing cali-
bration standards.
5.1.3 Preparation of Calibration Standards by Direct Gas Injection - This proce-
dure is applicable to organic compounds that exist entirely as a gas at ambient
conditions. The standards are prepared by direct injection of a known quantity of
a "pure" gas standard into a 10-liter Tedlar bag containing 5-0 liters of hydrocar-
bon-free air or nitrogen. If there is more than one target compound then multiple
component standards can be prepared by this method provided the relative elution
pattern for the compounds is known for the GC column being used. The following
procedures are used to prepare standards by direct gas injection:
1. Evacuate a previously leak checked, leakfree 10-liter Tedlar bag (also
checked for zero retention) equipped with a quick connect or equivalent
compatible to the connection on the Tedlar bag or the interface sample line
and preferably fitted with a septum-capped tee at the bag inlet (see Figure
5-5).
2. Fit a septum to the outlet of the gas cylinder containing the standard
component.
3. Meter 5-0 liters of hydrocarbon-free air or nitrogen into the bag at a rate
of 0.5 liter/min using a dry gas meter that has been calibrated in a manner
consistent with the procedure described in Subsection 2.1.2. At the start,
record dry gas meter pressure and temperature.
4. While the bag is filling, fill and purge a 0.5-ml gas-tight syringe with
the standard gas by withdrawing the gas from the cylinder through the
septum. Repeat the fill and purge of the syringe seven times before final-
ly filling the syringe and capping the needle with a GC septum. Allow the
syringe temperature to equilibrate with the ambient temperature.
5. Immediately before injecting the gas into the bag through the septum,
remove the septum cap, and adjust the syringe to the desired volume by
expelling the excess gas. The syringe should now be at ambient pressure.
Inject the gas into the bag through the septum (through the side of the bag
if it has not been fitted with a septum), withdraw the syringe, and imme-
diately cover any resulting hole with a piece of masking tape or the equi-
valent .
-------
Nitrogen
Cylinder
Gas
Tight
Syringo
Dry Gas Motor
Soptum
TedlarBag
Capacity
10 Liters
T) O Cfl
(3 p
-------
7.
8.
Section No. 3.16.5
Date June 30, 1988
Page 11
Record the final dry gas meter temperature and pressure, turn off the
dilution gas, and disconnect the Tedlar bag; record the ambient
temperature and pressure on a data form such as the one shown in Figure
5.6.
Place the bag on a smooth surface, and alternatively depress opposite
sides of the bag 50 times to mix the gases in the bag.
Calculate the organic standard concentration in the bag (Cs ) in ppmv using
the following formula.
x 106
293 x ps
T x 760
P. x T
Gv x 103 x
C_ =
293
Equation 5-5
y x
x 760
103
where
Gv = Gas volume of organic compound injected into the Tedlar bag,
ml,
106 = Conversion to ppmv, ul/liter,
Ps = Absolute pressure of syringe before injection, mm Hg,
Ts = Absolute temperature of the syringe before injection, °K,
Vm = Gas volume indicated by dry gas meter, liters,
y = Dry gas meter correction factor, dimensionless,
Pm = Average absolute pressure of the dry gas meter, mm Hg,
Tm = Average absolute temperature of dry gas meter, °K, and
lO3 = Conversion factor, ml/L.
Note: The syringe pressure and absolute temperature should equal the baro-
metric pressure and the absolute ambient temperature.
9. Prepare two more calibration standards sufficient to bracket the expected
concentration in the source samples.
10. Analyze the calibration standards following the procedures described in
Subsection 5-!•! for commercial gas cylinder mixtures. Use a pump on the
outlet side of the sample loop to flush the standards through the loop at
100 cc/min for 30 seconds. Using a manometer connected to a tee on the
outlet of the sample loop, make certain that the sample loop pressure
during analysis of the calibration standards is equal to the loop pressure
during actual sample analysis.
5.1.4 Preparation of Calibration Standards by Liquid Injection - This procedure is
used to prepare gaseous standards in Tedlar bags from liquid organic compounds.
The liquid compounds used must be 99 • 9% mole pure or the purity must be known to
calculate the gaseous standard concentrations. If there is more than one target
compound, then multiple component standards can be prepared by this method provided
the relative elution pattern for the compounds is known for the GC column being
used. Use the procedure that follows to prepare standards by this technique.
1. Assemble the equipment shown in Figure 5«7 using a dry gas meter calibrated
following the procedure described in Subsection 2.1.2 and a water manometer
for the pressure gauge. All connections,should be glass, Teflon, brass or
stainless steel with quick connects or equivalent, compatible to the con-
-------
Section No. 3.16.5
Date June 30, 1988
Page 12
Preparation of Standards in Tedlar Bags by Gas and Liquid Injection
o
Date:
Preparer:
Purpose: C*libr-e>hb*
held £$-/•
Organic Compound:
Compound Source:
Gas:
or Liquid:
Compouhd Purity (P) ;
Compound Mole Weight (M) : /£*>. 83
Gas Injection
Bag number or identification
Dry gas meter calibration factor ,(Y)
Final gas meter reading, liters
Initial gas meter reading, liters
Volume metered (Vm), liters
Ambient temperature, °C
Average gas meter temperature, °C
Absolute gas meter temp. (Tra), °K
Barometric pressure (Pb), mm Hg
Average gas meter pressure, mm Hg
Absolute gas meter press. (Pm), mm Hg
Gas volume injected (Gv), ml
Syringe temperature (T8), °K
Absolute syringe pressure (P ), mm Hg
Calculated concentration (Cs)
Mixture 1
T x P
1s x rra
v
Mixture 2
A//A-
Hixture 3
MM
s c a 1 c .
a c o r r .
x loo*
O
Liquid Injection
Bag number or identification
Dry gas meter calibration factor (Y)
Final gas meter reading* liters
Initial gas meter reading, liters
Volume metered (Vm ) , liters
Average gas meter temperature, °C
Absolute gas meter temp. (Tm), °K
Barometric pressure (Pb ) , mm Hg
Average gas meter pressure, mm Hg
Absolute gas meter press. (Pm), mm Hg
Liquid organic density (p) , ug/ml
Liquid volume injected (Lv), ul
Calculated concentration (Cs )
Mixture 1
S-l
17-.2&
Z-K
AS"
3->00
3-1,?-
Cs = 6.24 x 10* x
Lv x p x T0
Mixture 2
3-Z
1.10
7-.00
2,6
30!
. 4-
/.«#-
/.6.Z3
Z.00
J.4-
Mixture 3
S-3
O- 9430
ze.qo
2-6
75-6,4
/.I
2.00
/4-.S"
s caic
x 1002
M x Vm x y x Pm
Figure 5-6. Calibration data form for preparation of standards in Tedlar bags by
gas and liquid injection.
o
-------
SYRINGE
•N
s
If
NITROGEN CYLINDER
SEPTUM
MIDGET IMPINGER
-p:
TEDLAR BAG
, CAPACITY
50 LITERS
Figure 5-7- Apparatus for preparation of calibration standards by liquid
injection.
TJ O 03
to ta n>
03 rt o
(D (D ct
H-
»-» e-i O
W 3
OJ .
o
- OJ
VO O\
oo •
OOU1
-------
Section No. 3-16.5
Date June 30, 1988
Page 14
nection on the Tedlar bag or the interface sample line, for connection to
the Tedlar bag.
2. Allow the liquid organic compound to come to ambient temperature, and
determine the density of the liquid by weighing the liquid in a tare-
weighed ground-glass stoppered 25-ml volumetric flask or ground-glass
stoppered specific gravity bottle. Calculate the density in terms of
g/ml. As an alternative, use a literature value of the density of the
liquid at 20 °C.
3. Leak check the system by pressurizing it to 5 to 10 cm (2 to 4 in.) H20
and shutting off the diluent gas supply. The system is leakfree if there
is no change in the pressure after 30 seconds. If the leak check is
good, release the pressure. !lf the system fails the leak check, locate
the leak using a soap solution and repair the leak.
4. Connect a quick connect or equivalent, compatible to the connection on
the Tedlar bag or the interface sample line, to a leakfree uncontaminated
50-liter Tedlar bag. Evacuate the bag. " .
5. Turn on the hot plate and bring the water to a boil.
6. Connect the bag to the impinger outlet.
7. Record the initial meter reading and temperature. Open the diluent
supply valve, and adjust the flow rate to about 3 liters/minute so that
the bag will fill in about 15 minutes. Record the meter pressure and
temperature and the barometric pressure at the start on a form such as
the one shown in Figure 5-5-
8. Use a clean 1.0- or 10-microliter syringe with a needle of sufficient
length to inject the liquid below the air inlet branch of the tee on the
midget impinger.
9. Fill the syringe to the desired volume with the organic liquid, and
inject the liquid by inserting the needle through the septum until the
needle is below the air inlet. Depress the syringe plunger completely
over a period of about 10 seconds and withdraw the needle. NOTE: When
dispensing liquid from a syringe, take care to account for the volume of
liquid present in the syringe needle. In general, the potential error
resulting from the volume of the needle is most conveniently avoided by
ensuring that the needle volume is completely full of liquid upon
filling the syringe and dispensing from it. If air pockets exist in the
syringe after filling, this will be almost impossible.
10. When the bag is almost filled, record the water manometer pressure. Turn
off the diluent gas supply, and disconnect the bag. To equilibrate the
contents in the bag, either set the bag aside for an hour or massage the
bag by alternately depressing opposite sides of the bag 50 times.
11. Record the final meter reading and temperature. Calculate the con-
centration of the calibration standard (Cs) in the bag in ppmv using the
following formula.
— x p x 24.055 x 106
M Lv x p x Tn
Cs = = 6.24 x 10" x Equation 5-6
293 x Pm M x Vm x y x Pm
Vm x Y x x 103
Tm x 760
O
-------
Section No. 3-16.5
Date June 30, 1988
Page 15
where
Vra = Gas volume indicated by dry gas meter, liters,
Lv - Volume of liquid organic injected, ul,
M = Molecular weight of the organic compound, g/g-mole,
p = Organic liquid density, g/ml,
24.055 = Ideal molar gas volume at 293 °K and 760 mm Hg, liters/g-
mole,
106 = Conversion to ppmv, ul/liter,
103 = Conversion factor, ul/ml, and
y = Conversion factor for dry gas meter.
12. When using a liquid standard that is not 99-9# pure, use the following
formula to correct the calculated concentration of the calibration standard
(C« „„„) in ppm .
c al c
X P
Equation 5-7
where
s c or r
i
•'s c a 1 c
= Corrected calibration standard concentration, ppmv,
= Calculated calibration standard concentration (Cs), ppmv,
and
P = Purity of liquid organic compound, percent.
13. Prepare two more calibration standards sufficient to bracket the expected
concentration in the source samples.
14. Analyze the calibration standards following the procedures described in
Subsection 5»1«1 for commercial gas cylinder mixtures. Use a pump on the
outlet side of the sample loop to flush the standards through the loop at
100 cc/min for 30 seconds. Make certain that the sample loop pressure
during analysis of the calibration standards is equal to the loop pressure
during actual sample analysis.
An alternative procedure, subject to the approval of the Administrator, for
preparing gaseous standards from liquid organics substitutes a heated GC injection
port for the midget impinger setup and, due to the high back pressure of the injec-
tion port, a calibrated mass flowmeter for the dry gas meter.
5.1.5 Development of Relative Response Factors and Relative Retention Factors-
Far emission tests where on-site GC analysis involving more than one organic com-
pound will be conducted, the development and use of relative response factors and
relative retention times is recommended. In the laboratory, gaseous calibration
standards are prepared for each target organic compound and analyzed by one of the
techniques described in the previous subsections. Choose one of the target com-
pounds or prepare and generate another calibration curve for a different organic
compound to be used to calculate the relative response factors and retention times.
The compound selected should exhibit a retention time comparable to the other
target compounds, should be stable, and/or easy to prepare and use in the field.
This procedure must be verified in the laboratory prior to field testing.
The relative response factors are calculated by dividing the slopes of the
target compound calibration curve by the slope of the selected organic calibration
curve. The y-intercept from the regression equation is ignored in calculating the
relative response factors. It should be noted that a very large y-intercept
-------
Section No. 3.16.5
Date June 30, 1988
Page 16
(greater than 5% of the slope) for any compound may adversely affect the validity
of this calibration technique. During analysis of field samples, the selected
organic compound can be used to calibrate the GC detector response and column
performance. The response factor determined in the field for the selected organic
is used to calculate the field response factors for the other target compounds
using the relative response factors determined previously in the laboratory. The
same approach is used to predict the retention times of target compounds in the
field using the selected compound retention time determined in the field and the
relative retention times for the target compounds determined in the laboratory. Use
the following procedures to develop relative response factors and relative reten-
tion times.
1. Generate, at the minimum, a three-point calibration curve for each target
organic compound using gaseous standards following the procedures described
in the preceding subsections. Record the retention time of each compound.
2. Select one of the target compounds, preferably with a retention time be-
tween the other target compounds, or generate another calibration curve,
with a minimum of three points, for a non-target organic compound with a
comparable retention time. Select the standard compound to be used in the
field based primarily on the ease of use. Determine the retention time of
the selected compound (if not already determined). Measure the carrier gas
flow rate using a bubble-type flowmeter or other suitable flowmeter.
3. Inject a sample of the diluent gas, and determine the retention time of the
unretained diluent peak. This is needed to calculate the relative reten-
tion by the following formula:
* "R x i
rx/s - Equation 5-8
where
rx/« = Relative retention time based on adjusted retention volumes of the
target compounds and the selected compound, cc/cc,
tRxi = Initial retention time of compound x, seconds,
tM1 = Initial retention time of unretained diluent gas peak, seconds,
and
tRgl = Initial retention time of selected organic compound, seconds.
4. Calculate the relative response factor for each target compound relative to
the compound selected in step 2 using the following formula.
FRx = — Equation 5-9
where
FRx = Relative response factor for compound x, dimensionless,
ms = Slope from the calibration curve regression equation for the
selected organic compound, area counts/ppmv, and
mx = Slope from the calibration curve regression equation for
compound x, area counts/ppm .
o
/"""N
O
-------
Section No. 3-16.5
Date June 30, 1988
Page 1?
To verify that the relative response factors are correct, simulate the
transportation of the GC to the field by turning off the detector, the GC
oven and the carrier gas flow overnight or longer. After the simulated
period has elapsed, turn on the GC, the carrier gas, and the detector, and
establish the analytical conditions that were used to generate the relative
response factors. Measure the carrier gas flow rate.
Recalibrate the GC by generating a three-point calibration curve using the
selected organic compound, analyze each of the target compound calibration
standards, and a diluent gas sample. Calculate the concentration of each
target compound using the relative response factor for the compound and the
slope from the new calibration curve determined for the selected organic
compound with the following formula.
x F,
Rx
Equation 5-10
m
where
s t d
= Calculated concentration of compound x calibration standard,
m
s td
= Detector response for compound x calibration standard, area
counts,
= Slope from new calibration curve generated for selected
organic standard compound, area counts/ppmv, and
^"RX = Relative response factor for compound x, dimensionless.
The calculated value for each target compound using the compound's relative
response factor must be within 5% of the actual standard concentration for
this technique to be used for that compound.
Determine the predicted retention times for the target compounds using
their relative retention times and the retention time determined for the
selected organic compound using the following formula:
T x f
= ' '
T s t
M f
Equation 5"
where
Txf
?sf
Calculated retention time for compound x using the relative
retention time factor, seconds,
Measured retention for selected organic compound during
second analysis , seconds , and
tMf = Measured retention time of unretained diluent gas peak during
second analysis, seconds.
The calculated retention time for the target compounds should agree within
one second or 2%, whichever is greater, of the actual retention time seen
for the target compounds during the second analysis.
8. Record all data on a form such as the one shown in Figure 5-8.
5-1.6 Calibration Standards for Adsorption Tube Samples - The calibration stan-
dards necessary for the analysis of adsorption tube samples differ from the stan-
dards described in the previous subsections in that the adsorption tube standards
are liquid rather than gaseous. The liquid standards can be prepared directly in
the desorption solvent following the procedures described in the methods refer-
enced in Table F or, subject to the approval of the Administrator, on blank adsorp-
-------
. Section No. 3-16.5
Date June 30, 1988
Page 18
Development of Relative Response Factors and Relative Retention Factors
o
Date: ^-/Z-k /6& Preparer: \7- C?oo
Target Compound: Perch 1 or o eft** Lu/^J-.
Surrogate Compound: -Lsotyui^*^-
Target Compound Calibration Data
First analysis/verify analysis
Standard concentration .uL. '
Flow rate through loop (ml/min)
Liquid injection volume (tubes)
Injection time (24-hr clock)
Chart speed
Detector attenuation
Peak retention time (tRxi/tRxf)
Peak retention time range
Peak area
Peak area x atten. factor (Yi/Yx
Verification analysis conc.(Cx)
Percent deviation from actual
Calculated retention time (rTxf)
Percent deviation from actual
Linear regression equation; slope
Surrogate Calibration Data
First analysis/second analysis
Standard concentration
Flow rate through loop (ml/min)
Liquid injection volume (tubes)
Injection time (24-hr clock)
Chart speed
Detector attenuation
Peak retention time (tRsl/tRsf)
Peak retention time range
Peak area
Peak area x attenuation factor
fJrSc-1^- Purpose
Type of Standard
Type of Standard
Standard 1
_^.05>^
T.OO! tCQ
/V/f- / /l/4
)4'.CO/ /4:3'
/O&H/P'l*/ 10
Z. 1 -z-
/Z..43/ 2.,4-S
O.02-
39.3 / 34^
) 74O / (#2,
?. l^"
—l/,t%
2..4-Z
-0.07,
(mx): 73.38
Standard 1
/o-o,^
JLOO/^ 3JJO
f/A-l tifi
I4:oi/ /4-:2Lfc
ijfc / * / tD
1 1 }
'Z.ldl 3.2£>
O.O2-
7^3 / ^f
*7fc 3 / ?(**}
Linear regression equation; slope (m0 ) : <74-39 (m, ):
Nonretained peak retention time (tM1
Relative Response Factor (FRK): /. 2
/tnf): O.t\ /O.
•J- Relative Retei
: UUL Q>»™™*H«
: l-lquitl ,'r> B»Q
: C^ds Cif lt'ndJ&-
'
Standard 2
67.7-
2-OO / zoo
A/A-/ MAr
f5:t>*>/ /5".'4O
10 1 10
2. 1 2.
2.. 44 1 2, 4-5"
0.03
Wit/ /?&3
3 62^/39 2 4-
5-2.4-
+1.4
3..4-1-
O.(f7»
y-intercept (b)
Standard 2
SO-O^M
000/2^70
AJ fir/ fit A
/4,1 &/ /S'12,
/o //o
/ / 1
3.27/3.2.6
O.Ol
3-7&(o/ 3-W4-
!-•?(,&/ j-j-q 4-
&v*i't>L>(* rntxjiz
Standard 3
Z.QO! 2-QD
A/A- 1 A/-4
/4 .'/B//^:^
10/10
2-1 2-
Z.4-4/ 3,4-Z-
0.02.
?>(f52l $(e(*1e>
73o<-/7332-
-/.4
2.4-1.
0>(e"?t>
: lO.^tf
Standard 3
/CO ff>H\
300/2.00
/JAJ X/A
l^'All tb'&l
/O 1 /f
2.1 2-
3.11 / S.1(f
O-OI
4?l(p/4?£2-
1432./ VSO*
9£~./0 y-intercept (b) : -/6.0C
43
ntion Factor (rx/
,.): 0.1-
O
-x/s
'Txf
- tM.)
i - tM1)
x F,
Rx
m
s t d
f) x rx/s) + t,
"Mf
O
Figure 5-8. Data form for development of relative response and relative retention
factors.
'T>'.-,
-------
Section No. 3-16.5
Date June 30, 1988
Page 19
tion tubes and then desorbed. Both methods require similar preparation and analy-
sis of standards and desorption efficiency samples, but the way the calibration
curve is generated is different.
For calibration standards prepared directly in the desorption solvent, the
standards are used to generate the calibration curve, and the desorption efficiency
is determined separately. The calculated desorption efficiency is then used to
correct the analytical results for the emission test samples. The disadvantage of
this method is that the desorption efficiency may not be constant for each level.
This can result from a constant amount of analyte being retained by the adsorbent,
instead of an amount proportional to the total amount of analyte on the adsorbent.
When using the desorption efficiency to correct each analytical result, the analyst
must use the desorption efficiency determined for the concentration level closest
to that of the sample.
For calibration standards prepared on absorbent material, the desorbed solu-
tions are used to generate the calibration curve. By this procedure, the desorp-
tion efficiency is already taken into account when calculating the organic compound
catch of the adsorption tube samples. Liquid calibration standards must also be
prepared to calibrate the GC to determine if the desorption efficiency is greater
than 50#. The advantage of this method is that both level-dependent or absolute
amounts of organic compounds not desorbed from the adsorbent are automatically
taken into consideration.
For maximum accuracy, preparation of standards directly in the desorption
liquid or on adsorbent will require the preparation of a relatively large volume of
a high concentration working standard from which the calibration standards are
prepared. The working standard should be 100 times more concentrated than the
highest concentration calibration standard. Three levels of calibration standards
should be prepared to bracket the expected concentration of the liquid resulting
from desorption of actual samples. The concentration of the sample desorption
liquid will depend on the catch weight of the target organic compound(s) and the
amount of desorption liquid used (1.0 ml per 100 mg of adsorbent material). The
catch weight will in turn depend on the sample volume of flue or duct gas drawn
through the tubes and the concentration of the emission source. Use the following
formula to estimate the concentration (Cs), in ug/ml, of the midrange liquid stan-
dard that will be approximately equal to the actual samples:
Vpr«. X Cc X M
Cs = Equation 5-12
24.055 x Ld
where
Vpred = Predicted gas sample volume, liters,
Cc = Concentration of the organic compound at the source, ppmv
(ug-moles/g-mole),
M = Molecular weight of organic compound, ug/ug-mole,
24.055 = Ideal molar gas volume at 293 °K and 760 mm Hg, liters/g-mole, and
Ld = Volume of desorption liquid, ml.
The preliminary survey sample results should be used to calculate the required
calibration standard concentrations.
To prepare adsorption tube standards, use the procedure described in the refer-
enced method or the alternative procedure, subject to the prior approval of the
Administrator. Regardless of which type of calibration standard is selected, use
the following procedures to prepare the standards:
-------
Section No. 3.16.5
Date June 30, 1988
Page 20
1. Prepare a working standard for each organic compound by weighing each com-
pound into an individual tare-weighed ground-glass stoppered 250-inl volume-
tric flask. Dissolve the compound in the proper desorption solvent speci-
fied by the referenced method in Table E of the Method Highlights Section.
Dilute the solution to volume. Refrigerate the working standard when not
being used.
2. Using a clean microliter syringe, transfer the required amount of working
standard to a vial equipped with a Teflon-lined septum top, and add suffi-
cient desorption solvent to achieve a final volume equal to the volume of
desorption solvent required for actual samples. Cap the vial with the
top, shake the vial to mix the contents. NOTE: When dispensing liquid from
a syringe, take care to account for the volume of liquid present in the
syringe needle. In general, the potential error resulting from the volume
of the needle is most conveniently avoided by ensuring that the needle
volume is completely full of liquid upon filling the syringe and dispensing
from it. If air pockets exist in the syringe after filling, this will be
almost impossible.
3- Establish the optimum GC conditions determined during the analysis of the
preliminary survey samples.
4. Select a suitably sized injection syringe (5- or 10-ul), and flush the
syringe with acetone (or some other suitable solvent if acetone is the
standard component) to clean the syringe.
5. Flush the syringe with standard solution by withdrawing a syringe full of
the solution from the septum vial, and dispensing the solution into a
beaker containing charcoal adsorbent.
6. Refill the syringe with standard solution, withdraw the syringe from the
vial, and wipe the syringe needle with a laboratory tissue.
7. Adjust the syringe volume down to the desired amount (see NOTE under Step
2), and inject into the GC. Note the time of the injection on the strip
chart recorder and/or actuate the electronic integrator. Also, record the
standard concentration, detector attenuation factor, chart speed, injection
port temperature, column temperature and identity, and the carrier gas type
and flow rate on the form shown in Figure 5-9- It is also recommended that
the same information be recorded directly on the chromatogram. Record the
operating parameters for the particular detector being used.
8. After the analysis, determine the retention time of the standard component
and determine the peak area. Repeat the injection of the first liquid
standard until the area count from two consecutive injections yield area
counts within 5 percent of their average.
9. Multiply the average area count of the consecutive injections by the atten-
uation factor to get the calibration area value for that standard concen-
tration .
10. Repeat the procedure for the other standard concentrations.
11. Prepare a plot with the standard concentration (Cs) along the abscissa (x-
axis) versus the corresponding calibration area values along the ordinate
(y-axis). Perform a regression analysis to calculate the slope and the y-
intercept. Draw the least squares line on the plot.
To determine the desorption efficiency for the target organic compound(s)
requires spiking the target organic compound(s) onto the absorbent material and
desorbing the compound(s) using the same procedures that will be used for actual
samples; the desorption solution is then analyzed. The spikes should be prepared
at three levels in the range of the source samples. The following procedures are
used to determine the desorption efficiency:
o
o
o
-------
8
Section No. 3-16.5
Date June 30, 1988
Page 21
1. Place an amount of adsorbent material equivalent to the amount used for
actual tube samples in a vial with a Teflon-lined septum cap. Prepare ten
vials (three sets of triplicates and one blank).
2. Using a clean microliter syringe, aliquot from the working standard solu-
tion, in triplicate into each set of vials, an amount of spike equal to
each level of calibration standard.
3- Cap each vial immediately after spiking, and allow the vials to sit undi-
sturbed for the 30 minutes.
4. To desorb the ".spiked organic compound(s), dispense the appropriate volume
of desorbent solvent and treat the vials as specified by the referenced
method (Table E). Prepare a blank vial containing adsorbent and desorption
solvent only.
5- Analyze the desorption solutions following steps 4 through 8 used above for
the calibration standards. Record the data on the form shown in Figure
5-9.
6. Multiply the average area count of the consecutive injections by the atten-
uation factor to get the area value for that sample. NOTE: Attenuation
factors which affect the plot traced, but not the area count returned by
an electronic integrator should not be multiplied by the average area
count. Observe the effect of attenuation changes made at the console of a
specific electronic integrator to determine the appropriate course of
action.
7. If the desorption solutions are to be used to generate the calibration
curve, then plot the expected standard solution concentrations on the ab-
scissa (x-axis) and corresponding area value on the ordinate (y-axis).
Perform a regression analysis and draw the least squares line on the plot.
NOTE: If the desorption efficiencies of the selected solvent vary with
concentration for any of the organics to be analyzed, the relationship
between the expected standard solution concentrations and the corresponding
area value will not be strictly linear. Evaluate the linearity of the
resulting plot using control samples, and obtain the prior approval of the
Administrator before utilizing a least squares line generated from such
data.
Calculate the desorption efficiency (DE), in percent, for each level of
spike using the calibration area for the corresponding standard prepared
directly in the desorption solvent using the following formula:
A. - A_
DE =
x 100%
Equation 5-13
where
As = Average area value for desorption carried out at given concentration
level, area counts,
Ad = Average area value for desorption carried out on blank sample, area
counts, and
Ac = Average calibration area value for the corresponding standard level
prepared directly in the desorption solvent, area counts.
The desorption efficiency achieved at each level must be greater than 50% for
the adsorption tube sampling and analytical method to be acceptable. If adsorption
tubes have become the only remaining sampling option, and the 50% criteria cannot
-------
Section No. 3.16.5
Date June 30, 1988
Page 22
Preparation of Liquid Standards and Desorption Efficiency Samples
o
Date: Z/23/68 Preparer: &•
Organic Compound ;
Compound Source:
Compound Purity (P) ;
Liquid:
Compound Mole Weight (M) :
Adsorbent Material: Ar.tivb\td C&fat*^ Batch No: _/£]
Standards in Solvent ' Mixture 1
Desorption solvent volume (Va ) , ml 4-00
Compound spike amount (V0 ) , ul 4-°P
Organic compound density (p) , ug/ul /. £2-S
Standard concentration (C ) , ug/ml /. O2-
Standards on Adsorbent Mixture 1
Adsorbent amount, g £>.&W>
Compound spike amount (V0 ) , ul 4-. 00
Organic compound density (p) , ug/ul /, G>23
Desorption solvent volume {VB ) , ml 4. r>0
Desorption time, min. 3O
Standard concentration (C9 ) , ug/ml /•{?£-
o Desorption
Mixture 2
4.00
fi.OO
A 6>Z?
3. £5"
Solvent:
Mixture
4.00
I2.>0t>
1. 6>2J
/"/»,4f
Mixture 2 Mixture 3
0-8 CO
6.00
1. 62 i
4-.OD
5O
3.2.5"
0-800
f^CfO
J . (02$
4.0D
•30
6-.4T
W,
Blank
/.teZ3
4.00
GC Operating Conditions
Injection port temperature, °C
Carrier gas flow rate, ml/min
Column temperature:
Initial, °C
Program rate, °C/min
Final, °C
O
30
Chromatographic Results
Injection time, 24-hr clock
Distance to peak, cm
Chart speed, cm/min
Retention time, min
Attenuation factor
Standards in desorption solvent:
Peak area (Ac), area counts
Standards and blank from
adsorbent material:
Peak area (As and Ab),
area counts
Mixture 1 Mixture 2 Mixture 3 Blank
/5V 5?
47.3
/D
/O
4.73
.2-
857-
Desorption Efficiency Calculation
Desorption Efficiency (DE) , %
Mixture 1
-------
Section No. 3.16.5
Date June 30, 1988
Page 23
be met, then, subject to the prior approval of the Administrator, explore more
vigorous desorption techniques such as longer desorption times, Bonification of the
vials during desorption, and/or other desorption solvents.
5.2 Audit Sample Analysis
After analysis of the calibration standards, and generation of a calibration
curve, conduct the analysis of the audit cylinder(s). Audit. samples should be
introduced into the GC by the same procedure used for the calibration standards. If
possible, the audit sample should be introduced into the probe for the direct and
dilution interface techniques. The audit sample analysis must agree within 10% of
the actual concentration of the audit sample before sample analysis can begin. If
the audit criteria is not met, first try recalibrating the GC with the existing
standards, and then reanalyze the audit sample(s). If the 10% criteria still
cannot be met, remake the standards, recalibrate the GC, and reanalyze the audit
sample until the criteria is met or a representative of the Administrator decides
differently.
5.3 Sample Analysis
After the GC has been calibrated and the analysis of the audit sample(s) has
been conducted successfully, the samples can be analyzed. Use the same procedures
for sample analysis that were used to analyze the calibration standards. Record
the GC conditions and the analytical data on the form provided in Figure 5«1. The
following subse'ctions describe the procedures for analyzing Tedlai* bag samples,
direct and dilution interface samples, adsorption tube samples, and heated syringe
samples.
5.3-1 Analysis of Bag Samples - The following procedures are to be used to analyze
emission samples collected in Tedlar bags using a GC calibrated with gaseous cali-
bration standards prepared following one or more of the procedures described in
Subsection 5-1-
1. Attach a quick connect, or similar connecting device that is compatible
with the connection on the Tedlar bag to the gas sampling valve on the GC.
Attach A manometer connected to a tee on the outlet of the sample loop.
2. With the gas sampling valve in the load position, attach the first Tedlar
bag sample to the valve. Use a pump on the outlet side of the sample loop
to flush the sample through the loop at 100 cc/min for 30 seconds.
3. Turn off the pump, allow the sample loop to return to the same pressure
used during calibration standard analysis, and immediately switch the valve
to the inject position.
4. Note the time of the injection on the strip chart recorder and/or actuate
the electronic integrator. Also, record the sample identity, detector
attenuation factor, chart speed, sample loop temperature, column tempera-
ture and identity, and the carrier gas type and flow rate on a data form
such as Figure 5-1- It is also recommended that the game information be
recorded directly on the chromatogram. Record the operating parameters for
the particular detector being used.
5. Examine the chromatogram to ensure that adequate resolution is being
achieved for the major components of the sample. If adequate resolution is
not being achieved, vary the GC conditions until resolution is achieved,
and reanalyze the standards to recalibrate the GC at the new conditions.
6. After conducting the analysis with acceptable peak resolution, determine
the retention time of the sample components and compare them to the reten-
-------
Section No. 3-16.5
Date June 30, 1988
Page 24
tion times for the standard compounds. To qualitatively identify an indi-
vidual sample component as a target compound, the retention time for the
component must match within 0.5 seconds or 1%, whichever is greater, of the
retention time of the target compound determined with the calibration
standards.
7. Repeat the injection of the first sample until the area count for each
identified target compound from two consecutive injections give area counts
within 5 percent of their average.
8. Multiply the average area count of the consecutive injections by the atten-
uation factor to get the area value for that sample, and record the area
value on the data form provided in Figure 5.1. NOTE: When dispensing
liquid from a syringe, take care to account for the volume of liquid
present in the syringe needle. In general, the potential error resulting
from the volume of the needle is most conveniently avoided by ensuring
that the needle volume is completely full of liquid upon -filling the
syringe and dispensing from it. If air pockets exist in the syringe after
filling, this will be almost impossible.
9. Repeat the procedure for the other two samples collected at the same sampl-
ing location.
10. Immediately following the analysis of the last sample, reanalyze the cali-
bration standards, and compare the area values for each standard to the
corresponding area values from the first calibration analysis. If the
individual area values are within 5% of their mean value, use the mean
values to generate a final calibration curve for determining the sample
concentrations. If the individual values are not within 5# of their mean
values, generate a calibration curve using the results of the second analy-
sis of the calibration standards, and report the sample results compared to
both standard curves.
Determine the bag sample water content by measuring the temperature and the
barometric pressure near the bag. Use water saturation vapor pressure chart, assum-
ing the relative humidity of the bag to be 100% unless a lower value is known, to
determine the water vapor content as a decimal figure (% divided by 100). If the
bag has been heated during sampling, the flue gas or duct moisture content should
be determined using Method 4.
5-3«2 Analysis of Direct Interface Samples - Prior to analysis of the direct
interface sample, the GC should be calibrated using a set of gaseous standards
prepared by one of the techniques described in Subsection 5-1 and a successful
analysis of an audit sample should be completed. If possible, the audit samples
should be introduced directly into the probe. Otherwise, the audit samples are
introduced into the sample line immediately following the probe. The calibration
is done by disconnecting the sample line coming from the probe, from the gas sampl-
ing valve sample loop inlet, and connecting the calibration standards to the loop
for analysis. During the analysis of the calibration standards and the audit
sample(s), make certain that the sample loop pressure immediately prior to the
injection of the standards is at the same pressure that will be used for sample
analysis. To analyze the direct interface samples after GC calibration, use the
following procedures:
1. Reconnect the sample line to the inlet of the gas sample loop, switch the
valve to the load position, and turn on the sampling pump. Adjust the
sampling rate to at least 100 cc/minute, and, for the first sample, purge
the sample line long enough to flush the sample loop and the preceding
volume of tubing a minimum of 7 times.
o
o
o
-------
Section No. 3.16.5
Date June 30, 1988
Page 25
2. After purging the sampling system and the sample loop, decrease the sample
flow using the needle valve downstream of the loop until the loop pressure,
measured by a water manometer connected to a tee at the outlet of loop, is
equal to the pressure used during calibration.
3. Once the loop is at the correct pressure, immediately switch the sample
valve to the inject position. Note the time of the injection on the strip
chart recorder and/or actuate the electronic integrator. The flow through
the sample line can be returned to lOOcc/min after sample injection, and,
after the unretained compounds are detected, the gas sample valve can be
switched back to the load position. The system will then be ready to
inject the second sample as soon as the first analysis is completed.
4. Record the sample identity, detector attenuation factor, chart speed,
sample loop temperature, column temperature and identity, and the carrier
gas type and flow rate on a form such as Figure 5-1 •. It is also recom-
mended that the same information be recorded directly on the chromatogram.
Record the operating parameters for the particular detector being used.
5. Examine the chromatogram to ensure that adequate resolution is being
achieved for the major components of the sample. If adequate resolution is
not being achieved, vary the GC conditions until resolution is achieved,
and reanalyze the standards to recalibrate the GC at the new conditions.
6. Immediately after the first analysis is comple.te, repeat steps 2 and 3 to
begin the analysis of the second sample.
7. After conducting the analysis of the first sample with acceptable peak
resolution, determine the retention time of the sample components and
compare them to the retention times for the standard compounds. To quali-
tatively identify an individual sample component as a target compound, the
retention time for the component must match, within 0.5 seconds or 1%,
whichever is greater, the retention time of the target compound determined
with the calibration standards.
8. At the completion of the analysis of the second sample, determine if the
area counts for the two consecutive injections give area counts within 5
percent of their average. If this criterion cannot be met due to the
length of the analysis, and the emissions are known to vary because of a
cyclic or batch process, then the analysis results can still be used with
the prior approval of the Administrator.
9. Analyze a minimum of three samples collected by direct interface to consti-
tute an emissions test.
10. Immediately following the analysis of the last sample, reanalyze the cali-
bration standards, and compare the area values for each standard to the
corresponding .area values frpm the .first .calibration analysis. If the
individual area values are within 5% of their mean value.,, ,use the mean
values to generate a final calibration curve to determine the sample
concentrations. If the individual values are not within 5% of their mean
values, generate a calibration curve using the results of the second analy-
sis of the calibration standards, and report the sample results compared to
both standard curves.
5.3.3 Analysis of Dilution Interface Samples - For the analysis of dilution inter-
face samples, the procedures described for direct interface sampling in Subsection
5-3-2 should be followed, with the addition of a check of the dilution system.
-------
Section No. 3.16.5
Date June 30, 1988
Page 26
Prior to any sample analysis, the GC must first be calibrated, followed by the
dilution system check and an analysis of the audit sample(s). The audit sample(s)
are introduced preferably into the inlet to the dilution system or directly into
the gas sampling valve. Use the following procedures to conduct the check of the
dilution system:
1. Heat the dilution system to the desired temperature (0° to 3°C above the
source temperature) or, if the dilution system components can not tolerate
that temperature, to a temperature high enough to prevent condensation.
2. Adjust the dilution system to achieve the desired dilution rate, and intro-
duce a high concentration target gas into the inlet of the dilution system.
After dilution through the stage(s) to be used for actual samples, the
target gas should be at a concentration that is within the calibration
range.
3. Purge the gas sample loop with diluted high concentration target gas at a
rate of 100 cc/min for 1 minute, adjust the loop pressure measured by a
water manometer connected to a tee at the outlet of the loop, to the loop
pressure that was used during calibration and will be used during sample
analysis. The procedure for pressure adjustment for the sample loop will
vary with the type of dilution system that is used. In general, the loop
pressure can be lowered by reducing the flow into the loop and raised by
restricting the flow from the loop.
4. After achieving the proper loop pressure, immediately switch the gas sample
valve to the inject position.
5. Note the time of the injection on the strip chart recorder and/or actuate
the electronic integrator. Also, record the sample identity, detector
attenuation factor, chart speed, sample loop temperature, column tempera-
ture and identity, and the carrier gas type and flow rate on a form such as
Figure 5.1 • It is also recommended that the same information be recorded
directly on the chromatogram. Record the operating parameters for the
particular detector being used.
6. Determine the peak area and retention time for the target compound used for
the dilution check, and calculate the area value using the detector attenu-
ation. Compare the retention time to the retention time of the target
compound calibration standard. The retention times should agree within 0.5
seconds or 1%, whichever is greater. If the retention times do not agree,
identify the problem and repeat the dilution check.
7. Calculate the concentration of the dilution check gas (Cd) using the fol-
lowing formula.
Equation 5-14
S
where
Y = Dilution check target compound peak area, area counts,
b = y-intercept of the calibration curve, area counts,
S = Slope of the calibration curve, area counts/ppmv , and /"""N
d = Dilution rate of the dilution system, dimensionless. f J
-------
Section No. 3.16.5
Date June 30, 1988
Page 27
8. If the calculated value for the dilution check gas is not within 10# of the
actual dilution check gas, then determine if the GC or the dilution system
is in error. Check the calibration of the GC by analyzing one of the cali-
bration samples directly bypassing the dilution system. If the GC is
properly calibrated, then adjust the dilution system, and repeat the analy-
sis of the dilution check gas until the calculated results are within 1Q%
of the actual concentration.
Once the dilution system and the GC are operating properly, analyze the audit
sample(s). Upon completion of a successful audit, the system is ready to analyze
samples following the procedures described in Subsection 5-3-2. To load the sample
from the dilution system may not require a pump on the outlet of the sample loop,
but calibration of the GC using standards prepared in Tedlar bags will require a
pump. The system should be configured so that the pump can be taken off line when
it is not needed.
5-3«4 Analysis of Adsorption Tube Samples - Prior to the analysis of adsorption
tube samples, the target compounds adsorbed on the adsorption material must be
desorbed. The procedures found to give acceptable desorption efficiencies deter-
mined in Subsection 5.1.k should be used. The procedures for the analysis of the
sample desorption solutions are the same as those used for the standards. During
sample analysis, the sample collection efficiency must be determined. Use the
following procedures to determine the collection efficiency:
1. Desorb the primary and backup sections of the tubes separately using the
procedures found to give acceptable (50%) desorption efficiency for the
spiked adsorption material. Use the same final volume of desorption solu-
tion for the samples as was used for the standard solutions. If more than
one adsorption tube was used in series per test run, delay desorbing the
additional tubes until the analysis of the primary and backup section of
the first tube is complete, and the collection efficiency for the first
tube determined. Select the samples from the sampling run when the flue
gas or duct moisture was the highest and, if known, when the target com-
pound concentrations were the highest and analyze them first.
2. Calibrate the GC using standards prepared directly in desorption solvent or
prepared on adsorbent and desorbed.
3. Select a suitably sized injection syringe (5~ or 10-wl), and flush the
syringe with acetone (or some other suitable solvent if acetone is a target
compound) to clean the syringe.
4. Flush the syringe with the desorption solution from the tube's backup
section by withdrawing a syringe full of the solution from the septum vial,
and dispensing the solution into a beaker containing charcoal adsorbent.
5- Refill the syringe with the backup section desorption solution, withdraw
the syringe from the vial, and wipe the syringe needle with a laboratory
tissue.
6. Adjust the syringe volume down to the amount used for injecting standards
and inject the sample into the GC. Note the time of the injection on the
strip chart recorder and/or actuate the electronic integrator. Also,
record the sample identity, detector attenuation factor, chart speed,
injection port temperature, column temperature and identity, and the carri-
er gas type and flow rate on the data form shown in Figure 5•1• It is also
recommended that the same information be recorded directly on the chromato-
-------
Section No. 3.16.5
Date June 30, 1988
Page 28
gram. Record the operating parameters for the particular detector being
used.
7. After the analysis, determine the retention time of the major sample
components, and compare these retention times to the retention times deter-
mined for the target compounds during analysis of the standards. To quali-
tatively identify an individual sample component as a target compound, the
retention time for the component must match, within 0.5 seconds or 1%,
whichever is greater, the retention time of the target compound determined
with the calibration standards. Determine the peak area for each target
compound identified in the sample.
8. Repeat the injection of the first sample until the area counts for each
identified target compound from two consecutive injections are within 5
percent of their average.
9. Multiply the average area count of the consecutive injections by the atten-
uation factor to get the area value for that sample.
10. Next analyze the desorption solution from the primary section of the same
adsorption tube following steps 4 through 9 above.
11. For each target compound, calculate the total weight (W), in ug, present in
each section, taking into account the desorption efficiency using the
formula below.
(Y - b) 1
W or Wb = x — Equation 5-15
S DE
where
Y = Average value for the target compound in the section (primary or
backup), area counts,
b = y-intercept from the three-point calibration curve for the target
compound, area counts,
S = Slope from the three-point calibration curve for the target
compound, area/ug, and
DE = Desorption efficiency (if standards prepared directly in
desorption solvent are used for calibration).
12. Determine the percent of the total catch found in the primary section for
each target compound identified using the following formula.
mpx
Ecx = x 100# Equation 5-16
Kx + m*x)
where
ECJ£ = Collection efficiency of the primary section for target compound
x, percent,
mpx = Catch of compound x in the primary section, ug, and
mhx = Catch of compound x in the backup section, ug.
o
o
o
-------
Section No. 3.16.5
Date June 30, 1988
, Page 29
If the collection efficiency for the primary section for each target com-
pound identified is >^ 90%, then the collection efficiency for that compound
is acceptable. If the collection efficiency for all the target compounds
identified in the sample is acceptable, then the analysis of any additional
tubes used in series behind the first tube will not be necessary. Proceed
with the analysis of the other adsorption tube samples.
12. If the collection efficiency for any identified target compound is not
acceptable, then analyze the second tube (if used) connected in series and
determine the collection efficiency for that tube using the steps described
above. If the second tube does not exhibit acceptable collection and a
third tube was used, analyze the third tube. If acceptable collection
efficiency cannot be demonstrated for the sampling system, then the emis-
sion test using adsorption tubes will not be acceptable.
13. Immediately following the analysis of the last sample, reanalyze the cali-
bration standards, and compare the area values for each standard to the
corresponding area values from the first calibration analysis. If the
individual area values are within 5% of their mean value, use the mean
values to generate a final calibration curve for determining the sample
concentrations. If the individual values are not within 5% of their mean
values, generate a calibration curve using the results of the second analy-
sis of the calibration standards, and report the sample results compared to
both standard curves.
5.3-5 Analysts of Heated Syringe Gas Samples by Direct Injection - For the analy-
sis of samples collected in heated syringes, the GC will have to be equipped with
an injection septum fitted to the gas sampling valve sample loop inlet. Calibrate
the GC following one of the procedures described in Subsection 5.1 for gaseous
calibration standards. Analyze the heated syringe samples by the following proce-
dures :
1. Attach a GC septum to a quick connect, or equivalent, compatible with the
connector on the gas sampling valve, and attach this connector to the gas
sampling valve.
2. Insert the needle of the heated syringe through the septum, and purge the
sample loop by injecting a volume of the gas sample at least ten times
greater than the sample loop volume.
3. Allow the sample loop pressure, measured by a water manometer connected to
a tee on the outlet of the sample loop, to reach the same loop pressure
seen during analysis of the calibration standards, and immediately switch
the gas sample valve to the inject position.
4. Note the time of the injection on the strip chart recorder and/or actuate
the electronic integrator. Also, record the sample identity, detector
attenuation factor, chart speed, sample loop temperature and volume, column
temperature and identity, and the carrier gas type and flow rate on a form
such as Figure 5-1- It is also recommended that the same information be
recorded directly on the chromatogram. Record the operating parameters for
the particular detector being used.
5. Examine the chromatogram to ensure that adequate resolution is being
achieved for the major components of the sample. If adequate resolution is
not being achieved, vary the GC conditions until resolution is achieved,
and reanalyze the standards to recalibrate the GC at the new conditions.
-------
Section No. 3.16.5 x—v
Date June 30, 1988 ( )
Page 30 ^*~*/
6. After conducting the analysis with acceptable peak resolution, determine
the retention time of the sample components and compare them to the reten-
tion times for the standard compounds. To qualitatively identify an indi-
vidual sample component as a target compound, the retention tine for the
component must match, within 0.5 seconds or 1%, whichever is greater, the
retention time of the target compound determined with the calibration
standards.
7. Repeat the injection of the first sample until the area counts for each
identified target compound from two consecutive injections are within 5
percent of their average.
8. Multiply the average area count of the consecutive injections by the atten-
uation factor to get the area value for that sample.
9. Repeat the procedure for the other two samples collected at the same sampl-
ing location.
10. Immediately following the analysis of the last sample, reanalyze the cali-
bration standards, and compare the area values for each standard to the
corresponding area values from the first calibration analysis. If the
individual area values are within 5% of their mean value, use the mean
values to generate a final calibration curve for determining the sample
concentrations. If the individual values are not within 5% of their mean
values, generate a calibration curve using the results of the second analy-
sis of the calibration standards, and report the sample results compared to
both standard curves.
o
o
-------
Section No. 3.16.5
Date June 30, 1988
Page 31
Date Plant Name Sampling Location
Checks for Analysis of All Calibration Standards
A minimum of three concentration levels used for each target compound?
yes no. (The concentration used should bracket the expected
concentrations of the actual field samples.)
Proper GC conditions established prior to standard analysis? yes no.
(For initial conditions use analytical conditions found to be acceptable
during preliminary survey sample analysis.)
Individual peak areas for consecutive injections within 5% of their mean for
each target compound? yes no. (Repeat analysis of standards
until 5% criteria is met.)
Second analysis of standards after sample analysis completed? yes no.
Peak areas for repeat analysis of each standard within 5% of their mean peak
area? yes no. (If no, then report sample results compared to both
standard curves.)
Checks for Calibrations using Commercial Cylinder Gases
Vendor concentration verified by direct analysis? yes no.
Sample loop purged for 30 seconds at 100 ml/min prior to injection of calibra-
tion standards? yes no.
Checks for Preparation and Use of Calibration Standards Prepared by Dilution
Dilution system flowmeters calibrated? yes no. (Calibrate following
procedure described in Subsection 2.1.3.)
Sample loop purged for 30 seconds at 100 ml/min prior to injection of calibra-
tion standards? yes no.
Dilution ratio for dilution system verified? yes no. (Analysis of
low concentration cylinder gas after establishing calibration curve
recommended to verify dilution procedure, but not required since audit
sample will also verify dilution ratio.)
Figure 5-10. Postsampling operations checklist.
-------
Section No. 3.16.5
Date June 30, 1988 x—v
Page 32 (J
Figure 5.10 (Continued)
Checks for Preparation and Use of Calibration Standards by Direct Injection of
Gaseous Compounds or Liquid Injection
Tedlar bag used to contain prepared standard leak and contamination free?
yes no .
Dry gas meter used to fill bag calibrated? yes no. (Calibrate meter
following procedure described in Subsection 2.1.2.)
Organic standard material used for injection 99-9# pure? yes no. (If
no, then determine purity and use to correct calculated calibration
standard concentration.)
Prepared standard allowed to equilibrate prior to injection? yes no.
(Massage bag by alternately depressing opposite ends 50 times.)
Sample loop purged for 30 seconds at 100 ml/min prior to injection of calibra-
tion standards? yes no.
Development of Relative Response Factors and Retention Times
Suitable target organic or surrogate compound selected? yes no. I J
(Select compound that is stable, easy to prepare in the field, and has a
retention time similar to the target organic compounds.)
Relative response factors and retention times verified in the laboratory prior
to actual field use? yes no. (If no, verify following the
procedure described in Subsection 5.1.4.)
Checks for Preparation. Use, and Determination of Desorption Efficiency for Adsorp-
tion Tube Standards
Organic standard material used for injection 99.9# pure? yes no. (If
no, then determine purity and use to correct calculated calibration
standard concentration.)
Correct adsorbent material and desorption solvent selected? yes no.
(Refer to Table B in Method Highlights Section for proper adsorbent
material and desorption solvent.)
Desorption efficiency determined for adsorbent to be used for field sampling?
yes no. (If no, follow the procedure described in Subsection
5.1.5.)
O
-------
Section No. 3.16.5
Date June 30, 1988
Page 33
Figure 5.10 (Continued)
Checks for All GC Analysis of Field Samples
Check type of carrier gas used: helium , nitrogen , other
Carrier gas flow rate and pressure set correctly? yes no. (Carrier
gas flow rate and pressure set according to conditions developed during
presurvey sample analysis and within limitations of the GC as specified
by GC manufacturer.)
Oxygen and hydrogen flow rate and pressure for FID correct? yes no
(Oxygen and hydrogen gas flow rate and pressure for FID set according to
conditions developed during presurvey sample analysis and within
limitations of the GC as specified by GC manufacturer.)
Individual peak areas for consecutive injections within 5% of their mean for
each target compound? yes no. (Repeat analysis of standards
until 5# criteria is met.)
Audit sample analyzed and results within 10% of actual value? yes no.
(If no, recalibrate GC and/or reanalyze audit sample.)
Checks Type of Standard Used for Tedlar Bag Sample Analysis
Gas cylinders , dilution of gas cylinders , direct gas injection ,
direct liquid injection , and/or relative response factors and
retention times .
Checks For GC Analysis Of Tedlar Bag Samples
Sample loop purged for 30 sec. at 100 ml/min prior to injection of calibration
standards? yes no.
Stability of gas sample in Tedlar bag determined? yes no. (Deter-
mine stability by conducting a second analysis after the first at a time
period equal to the time between collection and the first analysis. The
change in concentration between the first and second analysis should be
less than 10%.)
Retention of target compounds in Tedlar bag determined? yes no. (If
no, then follow the procedure described in Subsection 5-3-1-)
Check GC Interface Technique Used
Direct Interface , 10:1 Dilution Interface , 100:1 Dilution Interface
-------
Section No. 3.16.5
Date June 30, 1988
Page 34
Figure 5.10 (Continued)
Checks For Suitability of GC Interface Technique
Analytical interference due to moisture content of source gas? _ yes _ no.
(Moisture in the source gas must not interfere with analysis in regard
to peak resolution according to EPA Method 625 criterion where the
baseline-to-valley height between adjacent peaks ic less than 2$% of the
sum of the two adjacent peaks.)
Physical requirements for equipment met on-site? _ yes _ no. (The
physical requirements for the equipment include sheltered environment,
"clean", uninterrupted power source suited for equipment, and adherence to
safety aspects related to explosion risk areas.)
Source gas concentration below level of GC detector saturation? _ yes _ no.
(Concentrations delivered to the detector can be reduced by using smaller
gas sample loops and/or dilution interface . )
Sampling systems purged with 7 changes of system volume prior to sample
analysis? _ yes _ no.
Check Type(s) of Standards Used for Interface Techniques
Gas cylinders _ , dilution of gas cylinders _ , direct gas injection _ ,
direct liquid injection _ , and/or relative response factors and
retention times _ .
Checks For Dilution Interface Analytical Apparatus
Dilution rate verified (within 10%) by introducing high concentration gas
through dilution system and analyzing diluted gas? _ yes _ no.
(If dilution rate not verified, then first check calibration of GC by
reanalyzing a calibration standard and then adjust dilution system to give
desired ratio) .
Sampling systems purged with 7 changes of system volume prior to sample
analysis? _ yes _ no.
Check Type of Standard Used for Adsorption Tube Analysis
Prepared directly in desorption solvent _ , and/or prepared on adsorbent and
desorbed _ .
Checks for GC Analysis of Adsorption Tube Samples
Desorption procedure used identical to procedure used to determine the
desorption efficiency? _ yes _ no.
S~\
o
o
-------
Section No. 3.16.5
Date June 30, 1988
Page 35
Figure 5-10 (Continued)
Collection efficiency determined for adsorption tubes used for actual field
sampling? yes no. (If no, then determine collection efficiency
following the procedures described in Subsection 5-3-4.)
Check Type of Standard Used for Analysis of Heated Syringe Samples
Gas cylinders , dilution of gas cylinders , direct gas injection ,
direct liquid injection , and/or relative response factors and
retention times
XT/ -•
-------
Section No. 3-16.5
Date June 30, 1988
Page 36
Table 5.1. ACTIVITY MATRIX FOR SAMPLE ANALYSIS
Characteristic
Calibration
Standards
All calibrations
Commercial gas
cylinder mixtures
Gas standards from
high concentration
gas cylinders
Standards prepared
by direct gas
injection
Standards prepared
by liquid
injection
(Continued)
Acceptance limits
1) Standard analysis
performed under same
GC conditions to be
used for samples
2) Three-point
(minimum) calibra-
tion curve generated
for each target
compound
3) Sufficient amount
of each standard to
recalibrate after
samples are analyzed
Certified by direct
analysis (within 5#
of manufacturer's
value); three levels
bracketing samples
Dilution ratio of
dilution system
verified (optional)
with calculated val-
ue using calibration
curve within 10%
of actual cone.
Gas injected 99.9#
pure, or calculated
standard concentra-
tion corrected for
gas impurity
Liquid injected
99-92 Pure, or
calculated standard
corrected for
liquid impurity
Frequency and method
of measurement
Before analysis of
calibration standards
determine sample ana-
lysis conditions
Before analysis ac-
quire or prepare stan-
dards for each target
compound at three
levels
Prior to initial
calibration and sample
analysis, determine
amount needed
Prior to use, check if
independent analysis
conducted and accept-
able and standards
will bracket samples
Prior to sample analy-
sis; calibration curve
from standards verif-
ied by analysis of an
undiluted sample
When calculating stan-
dard concentration,
determine purity of
gas standard
When calculating stan-
dard concentration,
determine purity of
liquid standard
Action if
requirements
are not met
Reanalyze stan-
dards under con-
ditions to be
used for samples
Acquire or pre-
pare standards at
at three levels
to bracket
samples
Acquire or
prepare
enough
standards
Procure certified
gas cylinders in
proper range
Identify and
correct problems
with dilution
system, and
remake, reana-
lyze, and re-
verify standards
Use pure gas or
determine purity
Use pure liquid
or determine
purity
o
o
o
-------
Table 5.1 (Continued)
Section No. 3.16.5
Date June 30, 1988
Page 37
Characteristic
Calibration
Standards
Relative response
factors and rela-
tive retention
times
Standards prepared
for adsorption
tube samples
Audit sample
analysis
Sample Analysis
All samples
Acceptance limits
Proper target or
surrogate standard
selected for on-site
calibration; method
verified (calculated
results within 10%
of actual concentra-
tion)
1) Liquid injected
99.9% pure, or
calculated standard
corrected for
2) Acceptable
desorption effici-
ency for target com-
pounds on adsorbent
material (>50#)
Analytical result
for audit sample
within 10% of actual
concentration
1) Audit sample
analysis within 10%
of actual cone.
2) Sample analysis
conditions the same
as conditions used
for analysis of
standards
Frequency and method
of measurement
When selecting stan-
dard choose stable,
easy to prepare stan-
dard with retention
time near or between
target compounds; ver-
fied following proced-
ures described in
Subsection 5.1.5
When calculating stan-
dard concentration,
determine purity of
liquid standard
During calibration
standard analysis
determine desorption
efficiency for each
target compound (see
Subsection 5.1.5)
After initial cali-
bration and prior to
sample analysis, ana-
lyze audit sample
Prior to sample ana-
lysis, analyze audit
sample
Prior to sample ana-
lysis check that ana-
lytical conditions are
the same as those used
for standard analysis
Action if
requirements
are not met
Select different
target or
surrogate
compound; if
procedure cannot
be verified
use calibration
standard for each
target compound
Use pure liquid
or determine
purity
Try longer de-
sorption times,
more vigorous
desorption condi-
tions , and/or
other desorbents
Reanalyze audit
sample, if not
acceptable, re-
make and reana-
lyze standards
Analyze audit
sample
Establish the
same analytical
conditions used
during analysis
of standards
(Continued)
-------
Table 5.1 (Continued)
Section No. 3-16.5
Date June 30, 1988
Page 38
o
Characteris tic
Sample Analysis
All samples
Bag samples
Acceptance limits
3) Retention times
for target compounds
identified in sample
within 0.5 seconds
or 1% of standards
4) Area counts for
consecutive injec-
tions of samples
within 5# of their
average for each
target compound
identified in sample
5) All three samples
constituting a test
analyzed together
6) After sample ana-
lysis, repeat analy-
sis of standards;
area counts for each
standard analysis
within 5% of their
mean
1) Bag sample moist-
ure content deter-
mined
2) Stability check
conducted on bag
content (<1Q% change
between first and
second analysis)
Frequency and method
of measurement
After analysis, deter-
mine retention times
for major components
in sample and compare
to standard retention
times
After second analysis
of a sample, calculate
average area for first
and second analysis
and percent difference
of single analysis
from the average
During sample analysis
After analysis of last
sample repeat standard
analysis; calculate
mean area counts and
percent difference for
each standard
During analysis using
vapor pressure chart
assuming 100% or known
value for relative
humidity
Second analysis con-
ducted n days after
first analysis where
n equals the number of
days between sample
collection and first
analysis
Action if
requirements
are not met
Qualitative
identification
requires reten-
times within 0.5
seconds or 1%;
repeat analysis
Repeat sample
injections until
consecutive in-
jections are
achieved meeting
the 5X criteria
for each target
compound
Analyze remaining
samples
Report sample
results using
both curves, if
5# criteria not
met
O
Measure ambient
pressure and
temperature near
bag
Conduct stabil-
ity check and if
criteria not met
then correct sam-
ple results with
approval of
Administrator
o
(Continued)
'
-------
Section No. 3.16.5
Date June 30, 1988
Page 39
Table 5.1 (Continued)
Characteristic
Sample Analysis
Direct interface
samples
Dilution interface
samples
Adsorption tube
samples
Acceptance limits
Two consecutive
injections give
area counts within
5# of their mean
1) Dilution ratio
verified (results
from analysis of
high concentration
standard through
dilution system
within 10% of actual
concentration
2) Two consecutive
injections give
area counts within
5% of their mean
Collection effici-
ency determined for
adsorption tubes
(902 of each target
compound identified
caught on primary
section)
Frequency and method
of measurement
After second analysis,
calculate average area
counts and percent
difference
Prior to sample ana-
lysis analyze high
concentration gas
introduced through
dilution system
After second analysis,
calculate average area
counts and percent
difference
Desorb and analyze
primary and backup
sections separately
Action if
requirements
are not met
Due to cyclic or
batch processes
and analysis
time, emission
levels may vary;
use results with
the prior appro-
val of the Ad-
ministrator
Identify problem;
recalibrate GC or
adjust dilution
system and repeat
analysis of high
concentration
gas
Due to cyclic or
batch processes
and analysis
time, emission
levels may vary;
use results with
the prior ap-
proval of the
Administrator
Analyze addi-
tional tube(s) if
used as backups
to first tube; if
criteria cannot
be met, test is
not valid
-------
o
o
o
-------
Section No. 3.16.6
Date June 30, 1988
Page 1
6.0 CALCULATIONS
Calculation errors due to procedural or mathematical mistakes can be a part of
total system error. Therefore, it is recommended that each set of calculations be
repeated or spotchecked, preferably by a team member other than the one who per-
formed the original calculations. If a difference greater than typical round-off
error is detected, the calculations should be checked step-by-step until the source
of error is found and corrected. A computer program is advantageous in reducing
calculation errors. If a standardized computer program is used, the original data
entered should be included in the printout so it can be reviewed; if differences
are observed, a new computer run should be made. Table 6.1 at the end of this
section summarizes the quality assurance activities for calculations.
Calculations should be carried out to at least one extra decimal figure beyond
that of the acquired data and should be rounded off after final calculation to two
significant digits for each run or sample. All rounding of numbers should be
performed in accordance with the ASTM 380-?6 procedures. All calculations should
then be recorded on a calculation form such as the ones in Figures 6.1 and 6.2 for
analysis by gas or liquid injection, respectively.
6.1 Calculations for GC Analysis Using Gas Injection
The same equation can be used to calculate the concentration of each organic
in Method 18 samples whenever the sampling technique used yields a gaseous sample
which can be injected into the GC. These techniques are: (1) the integrated bag
sampling technique, (2) the heated bag sampling technique, (3) the prefilled bag
sampling technique, (4) the direct interface sampling technique, and (5) the dilu-
tion interface sampling technique. This equation is used to calculate the sample
concentration (Cc) in ppm on a dry basis as follows:
Fr K
Equation 6-1
Pi Tr
where
B..
K
Concentration of organic from calibration curve, ppm,
Reference pressure, the barometric pressure or absolute sample loop
pressure recorded during calibration, mm Hg,
Sample loop temperature at time of sample analysis, °K,
Barometric or absolute sample loop pressure at time of sample analysis,
mm Hg,
Reference temperature, the temperature of the sample loop recorded
during calibration, °K,
Water vapor content of the stack gas, proportion by volume,
Relative response factor, if applicable (see Subsection 5-I-5). and
Dilution factor (applicable only for dilution interface and prefilled
bag sampling; for a 10 to 1 dilution, K = 10).
-------
Section No. 3.16.6
Date June 30, 1988
Page 2
6.2 Calculations for GC Analysis by Liquid Injection
For Method 18, liquid injection GC analyses are used in conjunction with the
adsorption tube sampling procedure. The same general equations are typically used
to calculate the concentration of each organic in a sample collected on an adsorp-
tion tube. However, the tester is referred to the National Institute of Occupa-
tional Health and Safety (NIOSH) method (see Table B in the Method Highlights
Section) for specifics on calculations for particular organics. The general equa-
tions are shown below.
6.2.1 Sample Volume Corrected to Standard Conditions on a Dry Basis - The correct-
ed sample volume (VBtd(dry) is calculated as shown.
T P v P V
Astd *bar vm „ „ *b a r vm
„ „
Vltd(d = - : - = 0.3858
P.«« T. (i -B../K)' T.
Equation 6-2
where
T,td/p,td = °-3858 °K/mm Hg,
Vn = Sample volume measured, L,
Pbar = Barometric pressure during sampling, mm Hg,
T8 = Temperature of sample gas, °K,
Bw$ = Water vapor of stack gas, proportion by volume, and
K = Dilution factor, if applicable.
"Note: Only apply thio correction if a denoicont is not used.
6.2.2 Desorptton Efficiency - Desorption efficiency (DE) for recovery of a speci-
fic compound using a certain solvent from an adsorption tube is calculated using
the following equation.
Qr -B
DE = - Equation 6-3
where
Qr = Average peak area for spiked tubes,
Qa = Average peak area for spiked solutions, and
B = Average peak area for media blanks.
6.3-3 Concentration of Organic in Sample - The concentration (C) of the organic in
the sample in milligrams per dry standard cubic meter or micrograms per dry stand-
ard liter (mg/dscm or ug/dsL) is calculated using the following equation.
o
-------
c =
(Wf + Wb - Bf - Bb)K
Section No. 3.16.6
Date June 30, 1988
Page 3
Equation 6-4
where
B.
K
V.td,
DE
dry
= Mass of organic found in primary sorbent section, ug,
= Mass of organic found in backup sorbent section, ug,
= Mass of organic found in primary section of average media
blank, ug,
= Mass of organic found in backup section of average media
blank, ug,
= Dilution factor, if applicable (for a 10 to 1 dilution, K = 10),
= Sample volume corrected to standard conditions and a dry
basis, L, and
= Desorption efficiency, decimal value.
6.2.4 Conversion to ppm - To convert the concentration in milligrams per dry
standard cubic meter (micrograms per dry standard liter) to ppm, the following
equation can be used.
ppn
where
C
MW
24.055 (dsL/g-mole gas) x C
MW
Concentration of organic, ug/dsL or mg/dscm, and
Molecular weight of organic, ug/ug-mole.
Equation 6-5
-------
SAMPLE CONCENTRATION
mm
K* = A/ A; . _, F/
Cs Pr Tt Fr K
*If applicable.
Section No. 3.16.6
Date June 30, 1988
Page *»
Equation 6-1
o
Figure 6.1. Calculation form for GC analysis by gas injection.
O
-------
jj Section No. 3.16.6
* Date June 30, 1988
Page 5
SAMPLE VOLUME, DRY BASIS AT STANDARD CONDITIONS
V,, = ^-^_ . ^ L, Pbar = _? JT^l • _2 mo Hg,
T = 3_ 0_ 0_ . D_ °K, Bws" =0._^_^, K*= _ _ •
P V
bar m i r~" ».
V.,, dpv = 0.3858 = £• & . 0 L Equation 6-2
" T. (1 -Bwg/K)" -
*If applicable.
DESORPTION EFFICIENCY
DE = (Qr - B)/Qa = 0 . __ _^ Equation 6-3
SAMPLE CONCENTRATION
wp - .82 A-V.VB. wb = j_ z_ o_ . 6_ ug, Bp = ___ g_ us,
Bb - ___ 0_ug, V8td = __^£".f.L, DE = O.f.£.
V8td
= 3 fj_ . J?_ og/dscm or ug/dsL Equation 6-4
CONVERSION TO PPM
C = ___ . _ mg/dscm or ug/dsL, MW = __ ^ _^ . _/ ug/ug-mole,
C „ = ' gas) x C = _ J_ £ . £ ppm Equation 6-5
MW
Figure 6.2. Calculation form for GC analysis by liquid injection.
-------
Section No. 3.16.6
Date June 30, 1988
Page 6
o
Table 6.1. ACTIVITY MATRIX FOR CALCULATION CHECKS
Characteristic
Analysis data
form
Calculations
Acceptance limits
All data and calcu-
tions are shown
Difference between
check and original
calculations should
not exceed round-off
error
Frequency and method
of measurement
Visually check
Repeat all calcula-
tions starting with
raw data for hand
calculations; check
all raw data input
for computer calcu-
lations; hand calcu-
late one sample per
test
Action if
requirements
are not met
Complete the
missing data
Indicate errors
on calculation
form. Figure 6.1
or 6.2
O
o
-------
Section No. 3.16.7
Date June 30, 1988
Page 1
7.0 MAINTENANCE
The normal use of emission- testing equipment subjects it to corrosive gases,
extremes in temperature, vibration, and shock. Keeping the equipment in good
operating order over an extended period of time requires knowledge of the equipment
and a program of routine maintenance which is performed quarterly or after 2830 L
(100 ft3) of operation, whichever is greater. In addition to the quarterly main-
tenance, a yearly cleaning of pumps and metering systems is recommended. Main-
tenance procedures for the various components are summarized in Table 7-1 at the
end of the section. The following procedures are not required, but are recommended
to increase the reliability of the equipment.
7.1 Pump
Several types of pumps may be used to perform Method 18; the two most common
are the fiber vane pump with in-line oiler and the diaphragm pump. The fiber vane
pump requires a periodic check of the oiler jar. Its contents should be translu-
cent; the oil should be changed if not translucent. Use the oil specified by the
manufacturer. If none is specified, use SAE-10 nondetergent oil. Whenever a fiber
vane pump starts to run erratically or during the yearly disassembly, the head
should be removed and the fiber vanes changed. Erratic operation of a diaphragm
pump is normally due to either a bad diaphragm (causing leakage) or to malfunctions
of the valves, which should be cleaned annually by complete disassembly.
7.2 Dry Gas Meter
Dry gas meters should be checked for excess oil or corrosion of the components
by removing the top plate every 3 months. Meters should be disassembled and all
components cleaned and checked whenever the rotation of the dials is erratic,
whenever the meter will not calibrate properly over the required flow rate range,
and during the yearly maintenance.
7.3 Rotameter
Rotameters should be disassembled and cleaned according to the manufacturer's
instructions using only recommended cleaning fluids every 3 months or upon erratic
operation .
7.4 Manometer
The fluid in the manometers should be changed whenever there is discoloration
or visible matter in the fluid, and during the yearly disassembly.
7-5 Sampling Train
All remaining sampling train components should be visually checked every 3
months and completely disassembled and cleaned or replaced yearly. Many items,
such as quick disconnects , should be replaced whenever damaged rather than checked
periodically. Normally, the best procedure for maintenance in the field is to have
on hand another entire unit such as a pump, Tedlar bags and containers, or heated
sample line rather than replacing individual components.
-------
Section No. 3.16.7
Date June 30, 1988
Page 2
7.6 Oas Chromatograph
o
Maintenance activities and schedules for gas chromatographs are make and nodol
specific. It is therefore recommended that the analyst consult the operator's
manual for instructions relative to maintenance practices and procedures.
O
o
-------
Section No. 3.16.7
Date June 30, 1988
Page 3
Table 7.1. ACTIVITY MATRIX FOR EQUIPMENT MAINTENANCE CHECKS
Apparatus
Fiber vane
pump
Diaphragm
pump
Dry gas meter
Rotameter
Manometer
Sampling
train
components
Gas chroina-
tograph
Acceptance limits
In-line oiler
free of leaks
Leak- free valves
functioning properly
No excess oil,
corrosion, or er-
ratic rotation of
the dial
Clean and no erra-
tic behavior
No discoloration or
visible matter in
the fluid
No damage
See owner's manual
Frequency and method
of measurement
Periodically check
oiler jar; remove
head and change fiber
vanes
Clean valves during
yearly disassembly
Check every 3 mo. for
excess oil or corro-
sion by removing the
top plate; check
valves and diaphragm
yearly and whenever
meter dial runs erra-
tically or whenever
meter will not cal-
ibrate
Clean every 3 no. or
whenever ball does
not move freely
Check periodically
and during disassemb-
ly
Visually check every
3 mo.; completely
disassemble and
clean or replace
yearly
See owner's manual
Action if require-
ments are not met
Replace as
needed
Replace when
leaking or mal-
functioning
Replace parts as
needed, or replace
meter
Replace
Replace parts
as needed
If failure noted,
replace appro-
priate components
See owner's manual
-------
o
o
o
-------
Section No. 3.16.8
Date June 30, 1988
Page 1
8.0 AUDITING PROCEDURES
An audit is an independent assessment of data quality. Independence is
achieved if the . individual(s) performing the audit and their standards and
equipment are different from the regular field team and their standards and
equipment. Routine quality assurance checks by a field team are necessary to
generate good quality data, but they are not part of the auditing procedure. Table
8.1 at the end of this section summarizes the quality assurance functions for
auditing.
Based on the requirements of Method 18 and the results of collaborative test-
ing of other Reference Methods, two specific performance audits are recommended:
1. An audit of the sampling and analysis of Method 18 is required for NSPS
and recommended for other purposes.
2. And audit of the data processing is recommended.
It is suggested that a systems audit be conducted as specified by the quality
assurance coordinator in addition to these performance audits. The two performance
audits and the systems audit are described in detail in Subsections 8.1 and 8.2,
respectively.
8.1 Performance Audits
Performance audits are conducted to evaluate quantitatively the quality of
data produced by the total measurement system (sample collection, sample analysis,
and data processing). It is required that cylinder gas performance audits be
performed once during every NSPS test utilizing Method 18 and it is recommended
that a cylinder gas audit be performed once during any enforcement source test
utilizing Method 18 conducted under regulations other than NSPS.
8.1.1 Performance Audit of the Field Test - As stated in Section 6.5 of 40 CFR 60,
Appendix A, Method 18, immediately after the preparation of the calibration curves
and prior to the sample analysis, the analysis audit described in 40 CFR 61, Appen-
dix C, Procedure 2: "Procedure for Field Auditing GC Analysis," should be per-
formed. The information required to document the analysis of the audit sample(s)
has been included on the example data sheets shown in Figures 8.1 and 8.2; the
complete text of the procedure is reproduced in Section 3.16.10. The audit anal-
yses shall agree within 10 percent (or other specified value, as explained below)
of the true value. When available, the tester may obtain audit cylinders by
contacting: U.S. Environmental Protection Agency, Atmospheric Research and Exposure
Assessment Laboratory, Quality Assurance Division (MD-77B), Research Triangle Park,
North Carolina 27711• Audit cylinders obtained from a commercial gas manufacturer
may be used provided that (i) the gas manufacturer certifies the audit cylinder in
a manner similar to the procedure described in 40 CFR 61, Appendix B, Method 106,
Section 5.2.3-It and (2) the gas manufacturer obtains an independent analysis.
Independent analysis is defined as an analysis performed by an individual other
than the individual who performs the gas manufacturer's analysis, while using
calibration standards and analysis equipment different from those used for the gas
manufacturer's analysis. Verification is completed and acceptable when the
independent analysis concentration is within 5 percent of the gas manufacturer's
concentration.
Responsibilities of the Audit Supervisor - The primary responsibilities of
the audit supervisor are to ensure that the proper audit gas cylinder (s) are or-
-------
Section No. 3.16.8
Date June 30, 1988
Page 2
dered and safe-guarded, and to interpret the results obtained by the analyst.
When auditing sampling systems that do not dilute the stack gases during samp-
ling, the audit gases ordered must consist of the same organic compound(s) that are
being tested; for emission standards on a concentration basis, the audit gas
concentration!s) must be in>tJierange of 25% to 2$Q% of the applicable standard. If
two cylinders are not available, then one cylinder can be used. If the audit
cylinder value is between 5 an^ 20 ppm, the agreement should be within 15 percent
of the stated audit cylinder value. It is strongly recommended that audit cylinder
values below 5 ppm not be used. For emission standards which specify a control
efficiency, the concentration of the audit gases should be in the range of 25% to
250/£ of the expected stack gas concentration. If two cylinders are not available,
the audit can be conducted using one cylinder.
The audit supervisor must ensure that the audit gas cylinder(s) are shipped to
the correct address, and to prevent vandalism, verify that they are stored in a
safe location both before and after the audit. Also, the audit cylinders should
not be analyzed when the pressure drops below 200 psi. The audit supervisor then
ensures that the audits are conducted as described below.
The audit supervisor must also interpret the audit results. When the measured
concentration agrees within 10 percent (or 15 percent for cylinders between 5 and
20 ppm) of the true value, he directs the analyst to begin analyzing the source
samples. When the measured concentration does not agree within the specified
criterion, the analyst should first recheck the analytical system and calculations,
and then repeat the audit. If the analyst fails the second audit, the audit
supervisor should have knowledge of the agency's policy for failure. If the
result(s) are close to the allowed percentage or a consistent bias is present, the
supervisor may wish to allow.the analyst use of a correction factor to be applied
at a later date; however, the analyst must make a significant effort to find the
discrepancy and correct it. If the error cannot be found, the audit supervisor
should allow analysis of the samples, and then conduct the audit again.
During the audit, the audit supervisor should record the appropriate cylinder
number(s), cylinder pressure(s) (at the end of the audit), and the calculated con-
centrations on the "Field audit report form", Figure 8.1. The individual being
audited must not, under any circumstances, be told the actual audit concentrations
until the calculated concentration^) have been submitted to the audit supervisor
and are considered acceptable.
When auditing sampling systems that dilute the emissions during collection,
the audit gas concentration value used in the calculations can either be based on
(1) the undiluted concentration using the criteria discussed above or (2) the
expected concentration of the gases following dilution during collection using the
same dilution factor as used for the emission samples.
The audit procedures that follow are presented according to the type of samp-
ling system used to collect the organic emissions and whether the samples are
analyzed on-site or at the base laboratory at a later date.
Container (Bag, Syringe, and Canister) Sampling with On-site Analysis - The
cylinder gas performance audit for rigid-container bag, syringe, or canister samp-
ling with on-site analysis consists of an on-site audit just prior to the analysis
of the emission samples. The recommended procedures for conducting the audit are
as follows:
1. The audit samples should be collected in the type of container that
will be used during the sample collection. However, to conserve on the
use of the audit gas(es), it is usually not necessary to use the rest
of the sampling system to collect the samples for unheated container
o
o
o
-------
Section No. 3.16.8
Date June 30, 1988
Page 3
FIELD AUDIT REPORT
Part A. - To be filled out by organization supplying audit cylinders.
1. Organization supplying audit sample(s) and shipping address
2. Audit supervisor, organization, and phone number
4.
5-
6.
Shipping instructions: Name, Address, Attention
/tettit Te-yh'ni. 100 Pnke Avf, SfevZ., /VC - X~.Af.
Guaranteed arrival date for cylinders -
Planned shipping date for cylinders -
Details on audit cylinders from last analysis
a. Date of last analysis. .....
b. Cylinder number
c. Cylinder pressure, psi
d. Audit gas (es) /balance gas..
e . Audit gas ( es ) , ppm
Low cone.
yW/C^gg
lOfrq.
IS'ltO
.£f^4.
...M....
High cone
• "z!%/*i
...M.....
Part B. - To be filled out by audit supervisor.
1. Process sampled
Audit location
2.
3-
4.
5-
Name of individual audit
Audit date
Audit Results:
b. Cylinder pressure before audit, psi
d. Measured concentration, ppm
Injection #1* Injection #2* Average
e. Actual audit concentration, ppm
f. Audit accuracy:1
Percent1 accuracy =
Measured Cone. - Actual Cone. x 100
Actual Cone.
Low
cone.
cylinder
/£&f
. y 5"fc£ .
'/s~p d '
J.4)/*+0
<*.**
»**£-
High
cone.
cylinder
/06&
-%£&-
2l2O/2,S4v
&?•
KM
1Results of two consecutive injections that meet the sample analysis
criteria of the test method.
Figure 8.1. Field audit report form.
-------
Section No. 3.16.8
Date June 30, 1988
Page 4
sampling. Problems related to the reaction or retention of the organic
compounds will still occur in the container. Other interferents in the
stack gas such as water vapor and other organics will not be present in
the audit cylinders and thus, related problems will not be assessed.
For heated container systems, it may be necessary to use the sampling
system to collect the audit gas. However, if the gases must be heated
to prevent condensation, it is likely that an audit gas cylinder will
not be available.
2. The audit samples should remain in the appropriate container approx-
imately the same length of time that the source samples will stay
prior to analysis. After! the preparation of the calibration curve, a
minimum of two consecutive analyses of each audit cylinder gas should
be conducted. The analyses must agree within 5# of the average. The
audit results should be calculated by the analyst (or representative)
and given to the audit supervisor. The audit supervisor will record
all the information and data on the"Field audit report form" and then
inform the analyst of the status of the audit. The equations for
calculation of error are included on the form.
Container (Bag and Canister) Sampling with Off-site Analysis - For cylinder gas
performance audits associated with rigid-container bag or canister samples that
are analyzed off-site, it is recommended that the audit be conducted off -site just
prior to the emission test (if the agency desires) and then repeated during the
off -site sample analysis as a quality control measure. The use of the pretest
audit will help ensure that the analytical system will be acceptable prior to
testing. Alternatively, the audit gas can be collected in the appropriate con-
tainer on-site or off -site, and then analyzed just prior to the analysis of the
field samples. It is recommended that the tester fill at least two containers with
the audit gas to guard against a leak causing a failed audit. Since the use of the
performance audit is to both assess and improve the data quality, the use of the
pretest audit will provide the tester /analyst with a better chance of obtaining
acceptable data. The recommended procedure for conducting the audit is the same as
above with the exception that the audit supervisor will likely not be present
during the audit and the data will be reported by telephone.
Direct Interface Sampling - Since direct interface sampling involves on-site
analysis , the performance audit is conducted on-site after the calibration of the
GC and prior to sampling. The audit gas cylinder is attached to the inlet of the
sampling probe. Two consecutive analyses of the audit gas must be within 5% of
the average of the two analyses. The tester /analyst then calculates the results
and informs the audit supervisor. The audit supervisor records all information and
results on the "Field audit report form" and then informs the tester/analyst as to
the acceptability of the results.
Dilution Interface Sampling - Since dilution interface sampling involves on-
site analysis, the performance audit is conducted on-site after the calibration of
the GC and prior to sampling. If the audit gas cylinder obtained has a concentra-
tion near the diluted sample concentration, the audit gas is introduced directly
into the sample port on the GC. If the audit gas cylinder obtained has a concen-
tration close to the expected sample concentration, then the audit gas is intro-
duced into the dilution system. The audit supervisor may wish to order one cylin-
der to assess both the dilution system and the analytical system and another cylin-
der to assess only the analytical system. Follow the same procedures described
/^\
o
-------
Section No. 3.16.8
Date June 30, 1988
Page 5
above for recording the information and reporting the results.
Adsorption Tube Sampling - The analysis for adsorption tube sampling is usually
conducted off-site. Therefore, the audit analysis is conducted off-site. Again,
the recommended procedure is to conduct the audit once prior to the test and again
following the test. Though the audit sample could be analyzed by direct
injection, the inclusion of the chromatogram printout in the report will prove that
the audit results were obtained through adsorption tube sampling and a solvent
extraction. Alternatively, the audit samples can be collected on-site or off-site
and then analyzed just prior to the analysis of the field samples. Since the audit
supervisor will likely not be present during the analysis, the results are reported
by telephone.
To collect the audit gas with the adsorption tube sampling train, connect a
sample "T" to the line from the audit gas cylinder. Place the adsorption tube
sampling system on one leg of the "T"; connect a rotameter to the other leg. With
the sampling system off, turn on the audit gas flow until the rotameter reads 2
1pm. Turn on the sampling system and sample the audit gas for the specified run
time. Approximately 1 1pm should be discharged through the rotameter.
8.1.2 Performance Audit of Data Processing - Calculation errors are prevalent in
processing data. Data processing errors can be determined by auditing the recorded
data on the field "and laboratory forms. The original and audit (check) calcula-
tions should agree within round-off error; if not, all of the remaining data should
be checked. The data processing may also be audited by providing the testing
laboratory with specific data sets (exactly as would appear in the field), and by
requesting that the data calculation be completed and that the results be returned
to the agency. This audit is useful in checking both computer programs and manual
methods of data processing.
8.2 Systems Audit
A systems audit is an on-site, qualitative inspection and review of the total
measurement system (sample collection, sample analysis, etc.). Initially, a
systems audit is recommended for each enforcement source test, defined here as a
series of three runs at one source. After the test team gains experience with the
method, the frequency of auditing may be reduced — for example, to once every four
tests.
The auditor should have extensive background experience in source sampling,
specifically with the measurement system being audited. The functions of the
auditor are summarized below:
1. Inform the testing team of the results of pretest audits, specifying any
area(s) that need special attention or improvement.
2. Observe procedures and techniques of the field team during sample collec-
tion.
3. Check/verify records of apparatus calibration checks and quality control
used in the laboratory analysis of control samples from previous source
tests, where applicable.
4. Record the results of the audit, and forward them with comments to the
test team management so that appropriate corrective action may be
initiated.
While on site, the auditor observes the source test team's overall perfor-
mance, including the following specific operations:
-------
Section No. 3.16.8
Date June 30, 1988
Page 6
1. Conducting the GC calibration and conducting the performance audit (if the
analysis is conducted on-site).
2. Setting up and leak testing the sampling train.
3. Collecting the sample at a proportional rate (if applicable) or constant
rate at the specified flow rate.
4. Conducting the final leak check and recovery of the samples.
5. Conducting the initial and final check on the dilution system (if appli-
cable) .
6. Sample documentation procedures, sample recovery, and preparation of
samples for shipment (if applicable).
7- Conducting sample analyses (if conducted on-site).
Figure 8.2 is a suggested checklist for the auditor.
o
o
o
-------
Section No. 3.16.8
Date June 30, 1988
Page 7
Yes
^
~7
S
,
,/
_k^
IX"
tX"
l/>
!/
^x
s
-VL.
IS
/
7?
No
^
><•
X
^
X
^
X
y/- /v
Comments
^
WA
$<*$ ftck
J^jL^^eoi
+-5-.S/0 %
'
fJ/A
A//A
N/L.
///A
rfffl
M/A
V/A
A//A-
X/A
0n -?/'/£
^ jj •. ^ . jj^ ^p X^
^^-s/Xe—^^
<^
^j/C
/^C/A /\^c* S"
Vjvjr ///yc^^c
Operation
PRESAMPLING PREPARATION
1. Knowledge of process operations
2. Results of pretest audit (+ 10% or other value)
3- Calibration of pertinent equipment, in
particular, dry gas meters and other flowmeters
4. Selection and checkout of equipment for proper
sampling and analytical techniques
BAGS - reactivity, condensation, & retention
ADSORPTION TUBES - adsorption & desorption
efficiency
DILUTION SYSTEM - dilution ratio
GC/COLUMN - adequate resolution
GC/DETECTOR - acceptable accuracy & precision
ON-SITE MEASUREMENTS
5- Results of on-site audit (+ 10% or other value)
6. Sampling system properly assembled
7. Based on pitot tube check, is proportional
sampling required (more than 10% flow change)
8. Dilution system check acceptable (if applicable)
9. Sampling system leak check acceptable
10. Proportional sampling properly conducted
11. Constant rate sampling properly conducted
12. Heater systems maintained at proper temperatures
13. Proper number of samples & sampling time
14. GC properly calibrated
15. Duplicate injections had acceptable precision <5%
16. Recording of pertinent process conditions during
sample collection, samples properly identified,
and calculations properly conducted
POSTSAMPLING
17. Results of off -site audit (+ 10% or other value)
18. GC properly calibrated
19. Duplicate injections had acceptable precision <5%
20. Adsorption efficiency accep table, > 90% on primary
21. Desorption efficiency acceptable, > 50% recovery
22. Adequate peak resolution
23. Bags passed reaction check, less than 10% change
24. Bags passed retention check, less than 5% retained
25. Flowmeters recalibration acceptable
26. Temperature sensor recalibration acceptable
COMMENTS
re0££.£e.£t fcbU .
Figure 8.2. Method 18 checklist to be used by auditors.
-------
Section No. 3.16.8
Date June 30, 1988
Page 8
Table 8.1. ACTIVITY MATRIX FOR AUDITING PROCEDURES
o
Apparatus
Performance
audit of
analytical phase
Data processing
errors
Systems audit—
observance
of technique
Acceptance limits
Measured relative
error of audit
samples less than
10% (or other stated
value) for both
samples
Original and checked
calculations agree
within round-off
error
Operational tech-
nique as described
in this section of
the Handbook
Frequency and method
of measurement
Frequency; Once during
every enforcement
source test*
Method; Measure audit
samples and compare
results to true values
Frequency; Once during
every enforcement
source test*
Method; Independent
calculations starting
with recorded data
Frequency; Once during
every enforcement
source test* until
experience gained,
then every fourth
test
Method; Observation of
techniques assisted
by audit checklist,
Figure 8.1
Action if
requirements
are not met
Review operating
technique and
repeat audit
Check and correct
all data for the
audit period
represented by
the sampled data
Explain to team
their deviations
from recommended
techniques and
note on Fig 8.1
O
*As defined here, a source test for enforcement of the NSPS comprises a series of
runs at one source. Source test for purposes other than enforcement of NSPS may
be audited at the frequency determined by the applicable group.
O
-------
Section No. 3.16.9
Date June 30, 1988
Page 1
9.0 RECOMMENDED STANDARDS FOR ESTABLISHING TRACEABILITY
To achieve data of desired quality, two essential considerations are
necessary: (1) the measurement process must bu in a state of statistical control
at the time of the measurement, and (2) the systematic errors, when combined with
the random variation (errors or measurment), must result in an acceptable
uncertainty. As evidence in support of good quality data, it is necessary to
perform qulaity control checks and independent audits of the measurement process;
to document these data; and to use materials, instruments, and measurement
procedures that can be traced to an apropriate standard of reference.
Data must be routinely obtained by repeat measurements of standar reference
samples (primary, secondary, and/or working standards) and the establishment of a
condition of process control. The working calibration standards should be
traceable to standards of higher accuracy.
Audit samples (as discussed in Section 3-16.8) must be used to validate test
results for compliance determination purposes and are recommendeed as an
independent check on the measurement process when the method is performed for other
purposes.
-------
o
o
o
-------
Section No. 3-16.10
Date June 30, 1988
Page 1
10.0 REFERENCE METHOD*
Since the initial promulgation of Method 18 in 1983 (48 FR, 48344 - 48360,
10/18/83), there have been a number of revisions and additions to the method. In
the interest of consistency and clarity, the version of Method 18 reproduced here
is from the most recent edition of the Code of Federal Regulations which
incorporates all promulgated changes to this date.
METHOD IB—MEASUREMENT or GASEOUS OR-
GAWIC CoMTOtmn EMISSIONS BV GAS CKRO-
MATOORAPHY
Introduction
This method should not be attempted by
persons unfamiliar with the performance
characteristics of gas chromatography. nor
by those persons who are unfamiliar with
source sampling. Particular care should be
exercised In the area of safety concerning
choice of equipment and operation In poten-
tially explosive atmospheres.
1. Applicability and Principle
1.1 Applicability. This method applies to
the analysis of approximately 00 percent of
the total gaseous organlcs emitted from an
Industrial source. It does not include tech-
niques to Identify and measure trace
amounts of organic compounds, such as
those found In building air and fugitive
emission sources.
This method will not determine com-
pounds that (1) are polymeric (high molecu-
lar weight). (2) can polymerize before analy-
sis, or (3) have very low vapor pressures at
stack or Instrument conditions.
1.2 Principle.
The major organic components of a gas
mixture are separated by gas chromatogra-
phy (GO and Individually quantified by
flame lonization, photolonization, electron
capture, or other appropriate detection
principles.
The retention times of each separated
component are compared with those of
known compounds under identical condi-
tions. Therefore, the analyst confirms the
identity and approximate concentrations of
the organic emission components before-
hand. With this Information, the analyst
then prepares or purchases commercially
available standard mixtures to calibrate the
GC under conditions identical to those of
the samples. The analyst also determines
the need for sample dilution to avoid detec-
tor saturation, gas stream filtration to elimi-
nate paniculate matter, and prevention of
moisture condensation.
2. Range and Sensitivity
2.1 Range. The range of this method Is
from about 1 part per million (ppm) to the
upper limit governed by GC detector satura-
tion or column overloading. The upper limit
can be extended by diluting the stack gases
with an inert gas or by using smaller gas
sampling loops.
2.2 Sensitivity. The sensitivity limit for a
compound Is defined as the minimum de-
tectable concentration of that compound, or
the concentration that produces a slgn&l-to-
noise ratio of three to one. The minimum
detectable concentration Is determined
during the presurvey calibration for each
compound.
3. Precision and Accuracy
Gas chromatographlc techniques typically
provide a precision of 5 to 10 percent rela-
tive standard deviation (RSD), but an expe-
rienced GC operator with a reliable instru-
ment can readily achieve 6 percent RSD.
For thla method, the following combined
GC/operator values are required.
(a) Precision. Duplicate analyses are
within 5 percent of their mean value.
(b) Accuracy. Analysis results of prepared
audit samples are within 10 percent of prep-
aration values.
4. Interferences
Resolution interferences that may occur
can be eliminated by appropriate GC
column and detector choice or by shifting
the retention times through changes In the
column flow rate and the use of tempera-
ture programming.
The analytical system is demonstrated to
b« essentially free from contaminants by pe-
riodically analyzing blanks that consist of
hydrocarbon-free air or nitrogen.
Sample cross-contamination that occurs
when high-level and low-level samples or
standards are analyzed alternately, is best
dealt with by thorough purging of the GC
sample loop between samples.
To assure consistent detector response,
calibration gases are contained In dry air.
To adjust gaseous organic concentrations
when water vapor is present In the sample.
water vapor concentrations are determined
for those samples, and a correction factor Is
applied.
S. Presurvey and Presurvey Sampling.
Perform a presurvey for each source to be
tested. Refer to Figure 18-1. Some of the in-
formation can be collected from literature
surveys and source personnel. Collect gas
samples that can be analyzed to confirm the
Identities and approximate concentrations
of the organic emissions.
5.1 Apparatus. This apparatus list also
applies to Sections 6 and 7.
'40 CFR 60, Appendix A, Method 18, July 1, 198?, pages ?40 - ?69-
-------
Section No. 3.16.10
Date June 30, 1988
Page 2
5.1.1 Tenon Tubing. (Mention of trade
names or specific products does not consti-
tute endorsement by the U.8. Environmen-
tal Protection Agency.) Diameter and
length determined by connection require-
ments of cylinder regulators and the OC.
Additional tubing Is necessary to connect
the OC sample loop to the sample.
5.1.2 Gas Chromatograph. GO with suit-
able detector, columns, temperature-con-
trolled sample loop and valve assembly, and
temperature programable oven. If necessary.
The GC shall achieve sensitivity require-
ments for the compounds under study.
5.1.3 Pump. Capable of pumping 100 ml/
mln. For flushing sample loop.
5.1.4 Flowmeters. To measure flow rates.
5.1.5 Regulators. Used on EOS cylinders
for GC and for cylinder standards.
5.1.6 Recorder. Recorder with linear strip
chart Is minimum acceptable. Integrator
(optional) Is recommended.
5.1.7 Syringes. 0.5-ml, 1.0- and 10-mlcro-
liter sizes, calibrated, maximum accuracy
(gas tight), for preparing calibration stand-
ards. Other appropriate sizes can be used.
5.1.8 Tubing Fittings. To plumb GC and
gas cylinders.
5.1.9 Septums. For syringe Injections.
5.1.10 Glass Jars. If necessary, clean-col-
ored glass Jars with Teflon-lined lids for
condensatj sample collection. Size depends
on volume of condensate.
5.1.11 Soap Film Flow Meter. To deter-
mine flow rates.
5.1.12 Tedlar Bags. 10- and 50-liter capac-
ity, for preparation of standards.
5.1.13 Dry Gas Meter with Temperature
and Pressure Gauges. Accurate to ±2 per-
cent, for perparatlon of gas standards.
5.1.14 Midget Implnger/Hot Plate As-
sembly. For preparation of gas standards.
5.1.15 Sample Flasks. For presurvey sam-
ples, must have gas-tight seals.
5.1.16 Adsorption Tubes. If necessary,
blank tubes filled with necessary adsorberd
(charcoal, Tenax, XAD-2, etc.) for presur-
vey samples.
5.1.17 Personnel Sampling Pump. Cali-
brated, for collecting adsorbent tube presur-
vey samples.
5.1.18 Dilution System. Calibrated, the
dilution system Is to be constructed follow-
ing the specifications of an acceptable
method.
5.1.19 Sample Probes. Pyrex or stainless
steel, of sufficient length to reach centrold
of stack, or a point no closer to the walls
than 1 m.
5.1.20 Barometer. To measure barometric
pressure.
5.2 Reagents.
5.2.1 Dslonized Distilled Water.
5.2.2 Methylene Dlchloride.
5.2.3 Calibration Gases. A series of stand-
ards prepared for every compound of inter-
eat.
5.2.4 Organic Compound Solutions. Pure
(09.0 percent), or as pure as can reasonably
be obtained, liquid samples of all the organ-
ic compounds needed to prepare calibration
standards.
5.2.5 Extraction Solvents. For extraction
of adsorbent tube samples In preparation
for analysis.
5.2.6 Fuel. As recommended by the man-
ufacturer for operation of the GC.
5.2.7 Carrier Gas. Hydrocarbon free, as
recommended by the manufacturer for op-
eration of the detector and compatability
with the column.
5.2.8 Zero Gas. Hydrocarbon free air or
nitrogen, to be used for dilutions, blank
preparation, and standard preparation.
5.3 Sampling.
8.3.1 Collection of Samples with Glass
Sampling Flasks. Presurvey samples can be
collected In precleaned 250-ml double-ended
glass sampling flasks. Teflon stopcocks,
without grease, are preferred. Flasks should
be cleaned as follows: Remove the stopcocks
from both ends of the flasks, and wipe the
parts to remove any grease. Clean the stop-
cocks, barrels, and receivers with methylene
dlchloride. Clean all glass ports with a soap
solution, then rinse with tap and delonlzed
distilled water. Place the flask In a cool
glass annealing furnace and apply heat up
to 500* C. Maintain at this temperature for
1 hour. Afte? this time period, shut off and
open the furnace to allow the flask to cool.
Grease the stopcocks with stopcock grease
end return them to the flask receivers.
Purse the assembly with high-purity nitrcX*^*\
sen for 2 to 5 minutes. Close off the stoi )
cocks after purging to maintain a slIjtiL J
positive nitrogen pressure. Secure the stop-*—'
cocks with tape.
Presurvey samples can be obtained either
by drawing the eases Into the previously
evacuated flask or by drawing the gases into
and purging the flask with a rubber suctlor.
bulb.
5.3.1.1 Evacuated Flask Procedure. Use a
high-vacuum pump to evacuate the flask to
the capacity of the pump; then close off the
stopcock leading to the pump. Attach a 6-
mm outside diameter (OD) glass tee to the
flask inlet with a short piece of Teflon
tubing. Select a 6-mm OD borosillcate sam-
pling probe, enlarged at one end to a 12-mm
OD and of sufficient length to reach the
centrold of the duct to be sampled. Insert a
glass wool plug In the enlarged end of the
probe to remove particulate matter. Attach
the other end of the probe to the tee with a
short piece of Teflon tubing. Connect a
rubber suction bulb to the third leg of the
tee. Place the filter end of the probe at the
centrold of the duct, or at a point no closer
to the walls than 1 m, and purge the probe
with the rubber suction bulb. After the
probe Is completely purged and filled with
duct eases, open the stopcock to the grab
flask until the pressure In the flask reaches
duct pressure. Close off the stopcock, and
remove the probe from the duct. Remove
the tee from the flask and tape the stop-
cocks to prevent leaks during shipment.
Measure and record the duct temperature
and pressure.
5.3.1.2 Purged Flask Procedure. Attach
one end of the sampling flask to a rubber
suction bulb. Attach the other end to a 6
mm OD glass probe as described in
-------
Section No. 3.16.10
Date June 30, 1988
Page 3
5.3.1.1. Place the filter end of the probe at
the centrold of the duct, or at a point no
closer to the walls than 1 m. and apply suc-
tion with the bulb to completely purge the
probe and flask. After the flask has been
purged, close off the stopcock near the suc-
tion bulb, and then close the stopcock near
the probe. Remove the probe from the duct,
and disconnect both the probe and suction
bulb. Tape the stopcocks to prevent leakage
during shipment. Measure and record the
duct temperature and pressure.
5.3.2 Flexible Bag Procedure. Tedlar or
aluminlzed Mylar bags can also be used to
obtain the presurvey sample. Use new bags.
and leak check them before field use. In ad-
dition, check the bag before use for con-
tamination by filling it with nitrogen or air,
and analyzing the gas by GC at high sensi-
tivity. Experience indicates that it is desira-
ble to allow the inert gas to remain in the
bag about 24 hours or longer to check for
desorption of organlcs from the bag. Follow
the leak check and sample collection proce-
dures given in Section 7.1.
5.3.3 Determination of Moisture Content.
For combustion or water-controlled process-
es, obtain the moisture content from plant
personnel or by measurement during the
presurvey. If the source is below 59' C,
measure the wet bulb and dry bulb tempera-
tures, and calculate the moisture content
using a psychrometric chart. At higher tern
peratures, use Method 4 to determine the
moisture content.
5.4 Determination of Static Pressure.
Obtain the static pressure from the plant
personnel or measurement. If a type S pltot
tube and an inclined manometer are used,
take care to align the pilot tube 90' from
the direction of the flow. Disconnect one of
the tubes to the manometer, and read the
static pressure; note whether the reading is
positive or negative.
5.5 Collection of Presurvey Samples with
Adsorption Tube. Follow Section 7.4 for pre-
survey sampling.
6. Analysis Development
6.1 Selection of GC Parameters.
6.1.1 Column Choice. Based on the initial
contact with plant personnel concerning the
plant process and the anticipated emissions,
choose a column that provides good resolu-
tion and rapid analysis time. The choice of
an appropriate column can be aided by a lit-
erature search, contact with manufacturers
of GC columns, and discussion with person-
nel at the emission source.
Most column manufacturers keep excel-
lent records of their products. Their techni-
cal service departments may be able to rec-
ommend appropriate columns and detector
type for separating the anticipated com-
pounds, and they may be able to provide In-
formation on Interferences, optimum oper-
ating conditions, and column limitations.
Plants with analytical laboratories may
also be able to provide information on ap-
propriate analytical procedures.
6.1.2 Preliminary GC Adjustment. Using
the standards and column obtained In Sec-
tion 6.1.1, perform initial tests to determine
appropriate GC conditions that provide
good resolution and minimum analysis time
for the compounds of interest.
6.1.3 Preparation of Presurvey Samples.
If the samples were collected on an adsorb-
ent, extract the sample as recommended by
the manufacturer for removal of the com-
pounds with a solvent suitable to the type
of QC analysis. Prepare other samples In an
appropriate manner.
6.1.4 Presurvey Sample Analysis. Before
analysis, heat the presurvey sample to the
duct temperature to vaporize any condensed
material. Analyze the samples by the GC
procedure, and compare the retention times
against those of the calibration samples
that contain the components expected to be
In the stream. If any compounds cannot be
Identified with certainty by this procedure.
Identify them by other means such as GC/
mass spectroscopy (GC/MS) or GC/lnfrared
techniques. A GC/MS system is recom-
mended.
Use the GC conditions determined by the
procedures of Section 6.1.2 for the first in-
jection. Vary the GC parameters during
subsequent Injections to determine the opti-
mum settings. Once the optimum settings
have been determined, perform repeat Injec-
tions of the sample to determine the reten-
tion time of each compound. To inject a
sample, draw sample through the loop at a
constant rate (100 ml/mln for 30 seconds).
Be careful not to pressurize the gas In the
loop. Turn off the pump and allow the gas
In the sample loop to come to ambient pres-
sure. Activate the sample valve, and record
Injection time, loop temperature, column
temperature, carrier flow rate, chart speed,
and attenuator setting. Calculate the reten-
tion time of each peak using the distance
from injection to the peak maximum divid-
ed by the chart speed. Retention times
should be repeatable within 0.5 seconds.
If the concentrations are too high for ap-
propriate detector response, a smaller
sample loop or dilutions may be used for gas
samples, and, for liquid samples, dilution
with solvent is appropriate. Use the stand-
ard curves (Section 6.3) to obtain an esti-
mate of the concentrations.
Identify all peaks by comparing the
known retention times of compounds ex-
pected to be in the retention times of peaks
in the sample. Identify any remaining un-
identified peaks which -have areas larger
than 5 percent of the total using a GC/MS,
or estimation of possible compounds by
their retention times compared to known
compounds, with confirmation by further
GC analysis.
6.2 Calibration Standards. Prepare or
obtain enough calibration standards so that
there are three different concentrations of
each organic compound expected to be
measured in the source sample. For each or-
ganic compound, select those concentrations
that bracket the concentrations expected in
the source samples. A calibration standard
may contain more than one organic com-
pound. If available, commercial cylinder
gases may be used if their concentrations
have been certified by direct analysis.
If samples are collected in adsorbent tubes
(charcoal, XAD-2, Tenax, etc.), prepare or
obtain standards In the same solvent used
for the sample extraction procedure. Refer
to Section 7.4.3.
-------
Section No. 3.16.10
Date June 30, 1988
Page 4
o
Verify the stability of all standards for
the time periods they are used. If gas stand-
ards are prepared in the laboratory, use one
or more of the following procedures.
6.2.1 Preparation of Standards from
High Concentration Cylinder Standards.
Obtain enough high concentration cylinder
standards to represent ail the organic com-
pounds expected in the. source samples.
Use these high concentration standards to
prepare lower concentration standards by
dilution, as shown by Figures 18-5 and 18-6.
To prepare the diluted calibration sam-
ples, calibrated rotameters are normally
used to meter both the high concentration
calibration gas and the diluent gas. Other
types of flowmeters and commercially avail-
able dilution systems can also be used.
Calibrate each flowmeter before use by
placing It between the diluent gas supply
and suitably sized bubble meter, splrometer.
or wet test meter. Record all data shown on
Figure 18-4. While it is desirable to calibrate
the cylinder gas flowmeter with cylinder
gas, the available quantity and cost may
preclude it. The error Introduced by using
the diluent gas for calibration is insignifi-
cant for gas mixtures of up to 1,000 to 2,000
ppm of each organic component.
Once the flowmeters are calibrated, con-
nect the flowmeters to the calibration and
diluent gas supplies using 0-mm Teflon
tubing. Connect the outlet side of the flow-
meters through a connector to a leak-free
Tedlar bag as shown in Figure 18-5. (See
Section 7.1 for bag leak-check procedures.)
Adjust the gas flow to provide the desired
dilution, and fill the bag with sufficient gas
for OC calibration. Be careful not to overfill
and cause the bag to apply additional pres-
sure on the dilution system. Record the flow
rates of both flowmeters, and the laborato-
ry temperature and atmospheric pressure.
Calculate the concentration C. In ppm of
each organic in the diluted gas as follows:
1C)
6
<> *
Eq. 18-1
where:
10'—Conversion to ppm.
X-Mole or volume fraction of the organic
in the calibration gas to be diluted.
q»-Flow rate of the calibration gas to be di-
luted.
q^x Diluent gas flow rate.
Single-stage dilutions should be used to pre-
pare calibration mixtures up to about 1:20
dilution factor.
For greater dilutions, a double dilution
system is recommended, as shown in Figure
18-6. Fill the Tedlar bag with the dilute gas
from the second stage. Record the laborato-
ry temperature, barometric pressure, and
static pressure readings. Correct the flow
reading for temperature and pressure. Cal-
culate the concentration C, in ppm of the
organic in the final gas mixture as follows:
Eq. 18-2
Where:
10*<> Conversion to ppm.
X-Mole or volume fraction of the organic
In the calibration gas to be diluted.
Ha -Flow rate of the calibration gas to be di-
luted in stage 1.
q^i-Flow rate of the calibration gas to be di-
luted In stage 2.
CU-Flow rate of diluent gas in stage 1.
dA-Flow rate of diluent gas In stage 2.
Further details of the calibration methods
for flowmeters and the dilution system can
be found in Citation 21 in the Bibliography.
6.2.2 Preparation of Standards from
Volatile Materials. Record all data shown on
Figure 18-3.
6.2.2.1 Oas Injection Technique. This
procedure b applicable to organic com-
pounds that exist entirely as a cas at ambi-
ent conditions. Evacuate a 10-liter Tedlar
bae that has passed a leak-check
-------
Section No. 3.16.10,,
Date June 30, 1988
Page 5 f
where:
G.-Gas volume or organic compound In-
jected, ml.
»•-Conversion to ppm.
P.-Absolute pressure of syringe before In-
Jectlon. mm He.
T.-Ateolute temperature of syringe before
Injection. '1C.
V.-Gas volume Indicated by dry gas meter.
liters.
Y-Dry gas meter calibration factor, dimen-
sionleEs.
P.-Absolute pressure of dry gas meter, mm
Hg.
T.-Absolute temperature of dry gas meter.
•K.
1000-Conversion factor, ml/liter.
6.2.2.2 Liquid Injection Technique. Use
the equipment shown in Figure 18-8. Cali-
brate the dry gas meter as described in Sec-
tion 6.2.2.1 with a wet test meter or a spl-
rometer. Use a water manometer for the
pressure gauge and glass. Teflon, braes, or
stainless steel for all connections. Connect a
valve to the inlet of the SO-llter Tedlar bag.
To prepare the standards, assemble the
equipment as shown in Figure 18-8, and
leak-check the system. Completely evacuate
the bag. Fill the bag with hydrocarbon-free
air, and evacuate the bag again. Close the
inlet valve.
Turn on the hot plate, and allow the
,ter to reach boiling, Connect the bag to
s impinger outlet. Record the Initial
'eter reading, open the bag inlet valve, and
open the cylinder. Adjust the rate so that
the bag will be completely filled in approxi-
mately 15 minutes.'Record meter pressure
and temperature, and local barometric pres-
sure.
Allow the liquid organic to equilibrate to
room temperature. Fill the 1.0- or 10-micro-
liter syringe to the desired liquid volume
with the organic. Place the syringe needle
Into the impinger inlet using the septum
provided, and Inject the liquid Into the flow-
Ing air stream. Use a needle of sufficient
length to permit Injection of the liquid
below the air inlet branch of the tee.
Remove the syringe.
When the bag is filled, stop the pump, and
close the bag inlet valve. Record the final
meter reading, temperature, and pressure.
Disconnect the bag from the Impinger
outlet, and either set it aside for at least 1
hour, or massage the bag to Insure complete
mixing.
Measure the solvent liquid density at
room temperature by accurately weighing a
known volume of the material on an analyt-
ical balance to the nearest 1.0 milligram. A
ground-glass stoppered 25-mil volumetric
flask or a glass-stoppered specific gravity
bottle is suitable for weighing. Calculate the
result in terms of g/ml. As an alternative,
literature values of the density of the liquid
at 20 *C may be used.
Calculate each organic standard concen-
tration C. in ppm as follows:
Lv D (24.055 x 106)
M
100°
= 6.24 x 10
Y P
Eq. 18-4
where:
L,-Uquid volume of organic injected, pi.
pi-Liquid organic density as determined, E/
ml.
M« Molecular weight of organic, g/g-mole.
24.055-Ideal gas molar volume at 293 *K
and 760 mm He, liters/g-mole.
101-Conversion to ppm.
1CCO-Conversion factor, jd/mL
6.3 Preparation of Calibration Curves.
Establish proper GC conditions, then flush
the campling loop for 30 seconds at a rate of
100 ml/mln. Allow the sample loop pressure
to equilibrate to atmospheric pressure, and
activate the injection valve. Record the
standard concentration, attenuator factor,
injection time, chart speed, retention time.
peak area, sample loop temperature, column
temperature, and carrier gas flow rate.
Repeat the standard Injection until two con-
secutive Injections give area counts within S
percent of their average. The average value
multipled by the attenuator factor is then
the calibration area value for the concentra-
tion.
Repeat this procedure for each standard.
Prepare a graphical plot of concentration
(C,) versus the calibration area values. Per-
form a regression analysis, and draw the
least squares line.
6.4 Relative Response Factors. The cali-
bration curve generated from the standards
for a single organic can usually be related to
each of the Individual GC response curves
that are developed In the laboratory for all
the compounds In the source. In the field,
standards for that single organic can then
be used to "calibrate" the GC for all the or-
ganics present. This procedure should first
be confirmed in the laboratory by preparing
and analyzing calibration standards contain-
ing multiple organic compounds.
6.5 Quality Assurance for Laboratory
Procedures. Immediately after the prepara-
tion of the calibration curves and prior to
the presurvey sample analysis, the analysis
audit described In 40 CFR Part 61. Appen-
dix C, Procedure 2: "Procedure for Field Au-
diting OC Analysis," should be performed.
The information required to document the
analysis of the audit samples has been In-
cluded on the example data sheets shown in
Figures 18-3 and 18-7. The audit analyses
should agree with the audit concentrations
within 10 percent. When available, the
tester may obtain audit cylinders by con-
tacting: U.S. Environmental Protection
Agency, Environmental Monitoring Systems
-------
Laboratory. Quality Assurance Division
(MD-77), Research Triangle Park, North
Carolina 27711. Audit cylinders obtained
from a commercial gas manufacturer may
be used provided that (a) the gas manufac-
turer certifies the audit cylinder In a
manner similar to the procedure described
In 40 CFR Part 61. Appendix B, Method 108.
Section 5.2.3.1, end (b) the gas manufactur-
er obtains an Independent analysis of the
audit cylinders to verify this analysis. Inde-
pendent analysis Is defined as an analysis
performed by an Individual other than the
individual who performs the eta manufac-
turer's analysis, while using calibration
standards and analysis equipment different
from those used for the gas manufacturer's
analysis. Verification is complete and ac-
ceptable when the Independent analysis
concentration is within S percent of the fas
manufacturer's concentration.
7. Final Sampling and Analytit Procedure
Considering safety (flams hessrds) and
the source conditions, select an appropriate
sampling and analysis procedure (Section
7.1,7.2,7.3, or 7.4). In situations where & hy-
drogen flame la a hazard and no intrinsical-
ly safe OC is suitable, use the flexible bag
collection technique or an adsorption tech-
nique. It the source temperature is below
100'C, and the organic concentrations are
suitable for the detector to be used, use the
direct Interface method. If the source cases
require dilution, use a dilution Interface and
either the bag sample or adsorption tubes.
The choice between these two techniques
will depend on the physical layout of the
site, the source temperature, and the stor-
age stability of the compounds If collected
in the bag. Sample polar compounds by
direct interfacing or dilution Interfacing to
prevent sample loss by adsorption on the
bag.
7.1 Integrated Bag Sampling and Analy-
sis.
7.1.1 Evacuated Container Sampling Pro-
cedure. In this procedure, the bags are filled
by evacuating the rigid air-tight containers
that hold the bags. Uss a field sample data
sheet as shown In Figure 18-10. Collect trip-
licate sample from each sample location.
7.1.1.1 Apparatus.
7.1.1.1.1 Probe. Stainless steel, Pyrex
glass, or Teflon tubing probe, according to
the duet temperature, with 6.4-mm OD
Teflon tubing of sufficient length to con-
nect to the sample bag. Use stainless steel or
Teflon unions to connect probe and sample
line.
7.1.1.1.2 Quick Connects. Male (2) and
female (2) of stainless steel construction.
7.1.1.1.3 Needle Valve. To control CM
flow.
7.1.1.1.4 Pump. Leakless Teflon-coated
diaphragm-type pump or equivalent. To de-
liver at least 1 liter/mln.
7.1.1.1.6 Charcoal Adsorption Tube. Tube
filled with activated charcoal, with glass
wool plugs at each end, to adsorb organic
vapors.
7.1.1.1.6 Flowmet«r. 0 to 800-ml flow
range: with manufacturer's calibration
curve.
Section No. 3.16.10
Date June 30, 1988
Page 6
7.1.1.2 Sampling Procedure. To obtain a
sample, assemble the sample train as shown
in Figure 18-0. Leak check both the bag and
the container. Connect the vacuum line
from the needle valve to the Teflon sample
line from the probe. Place the end of the
probe at the centrold of the stack, or at a
point no closer to the walls than 1 m, and
start the pump with the needle valve adjust-
ed to yield a flow of 0.5 liter/minute. After
allowing sufficient time to purge the line
several times, connect the vacuum line to
the bag, and evccuate until the rotamcter
Indicates no flow. Then position the sample
and vacuum lines for sampling, and begin
th« actual sampling, keeping the rate pro-
portional to the Bluet velocity. As a precau-
tion, direct the gas exiting the rotameter
away from sampling personnel. At the end
of the sample period, shut off the pump,
disconnect the sample line from the bag,
and disconnect the vacuum line from the
bag container. Record the source tempera-
ture, barometric pressure, ambient tempera-
ture, sampling flow rate, and Initial and
final sampling time on the data sheet shown
in Figure 18-10. Protect the Tedlar bag and
Ito container from sunlight. When possible,
perform the analysis within 2 hours of
sample collection.
7.1.2 Direct Pump Sampling Procedure.
Follow 7.1.1, except place the pump and
needle valve between the probe and the bag.
Ues a pump and needle valve constructed of
stainless steel or some other material not af-
fected by the stack gas. Leak check the
system, end then purge with stack gas
before the connecting to the previously
evacuated bag.
7.1.3 Explosion Risk Area Bag Sampling
Procedure. Follow 7.1.1 except replace the
pump with another evacuated can (see
Figure 18-Oa). Use this method whenever
there Is a possibility of an explosion due to
pumps, heated probes, or other flame pro-
ducing equipment.
7.1.4 Other Modified Bag Sampling Pro-
cedures. In the event that condensation is
observed in the bag while collecting the
sample and a direct interface system cannot
be used, heat the bag during collection, and
maintain It at a suitably elevated tempera-
ture during all subsequent operations.
(Note: Take care to leak check the system
prior to the dilutions so as not to create a
potentially explosive atmosphere.) As an al-
ternative, collect the sample gas, and simul-
taneously dilute It in the Tedlar bag.
In the first procedure, heat the box con-
taining the sample bag to the source tem-
perature, provided the components of the
bag and the surrounding box can withstand
this temperature. Then transport the bag as
rapidly as possible to the analytical area
while maintaining the heating, or cover the
box with an Insulating blanket. In the ana-
lytical area, keep the box heated to source
temperature until analysis. Be sure that the
method of heating the box and the control
for the heating circuit are compatible with
the safety restrictions required in each area.
o
o
o
-------
Section No. 3-16.10
Date June 30, 1988
Page 7
To use the second procedure, preflll the
Tedlar bag with a known quantity of Inert
etas. Meter the Inert gas Into the bag accord-
Ing to the procedure for the preparation of
gas concentration standards of volatile
liquid materials (Section 6.2.2.2). but elimi-
nate the midget Impincer section. Take the
partly filled bag to the source, and meter
the source gas Into the bag through heated
sampling lines and a heated flowmeter. er
Tenon positive displacement pump. Verify
the dilution factors periodically through di-
lution and analysis of gases of known con-
centration.
7.1.5 Analysis Of Bag Samples.
7.1.5.1 Apparatus. Same as Section 5. A
minimum of three gas standards are re-
quired.
7.1.5.2 Procedure. Establish proper OC
operating conditions as described in Section
6.3. and record all data listed in Figure 18-7.
Prepare the GC so that gas can be drawn
through the sample valve. Flush the sample
loop with gas from one of the three calibra-
tion mixtures, and activate the valve.
Obtain at least two chromatograms for the
mixture. The results are acceptable when
the peak areas from two consecutive Injec-
tions agree to within 5 percent of their aver-
age. If they do not, run additional analyses
or correct the analytical techniques until
this requirement is met. Then analyze the
other two calibration mixtures in the same
manner. Prepare a calibration curve as de-
scribed In the same manner. Prepare a cali-
bration curve as described In Section 6.3.
Analyze the source gas samples by con-
necting each bag to the sampling valve with
a piece of Tenon tubing Identified for that
bag. Follow the specifications on replicate
analyses specified for the calibration gases.
Record the data listed in Figure 18-11. If
certain Items do not apply, use the notation
"N,A." After all samples have been ana-
lyzed, repeat the analyses of the calibration
gas mixtures, and generate a second calibra-
tion curve. Use an average of the two curves
to determine the sample gas concentrations.
If the two calibration curves differ by more
than 5 percent from their mean value, then
report the final results by comparison to
both calibration curves.
7.1.6 Determination of Bag Water Vapor
Content. Measure and record the ambient
temperature and barometric pressure near
the bag. From a water saturation vapor
pressure table, determine and record the
water vapor content as a decimal figure.
(Assume the relative humidity to be 100 per-
cent unless a lesser value Is known.) If the
bag has been maintained at an elevated tem-
perature as described In Section 7.1.4, deter-
mine the stack gas water content by Method
4.
7.1.7 Quality Assurance. Immediately
prior to the analysis of the stack gas sam-
ples, perform audit analyses as described In
Section 6.5. The audit analyses must agree
with the audit concentrations within 10 per-
cent. If the results are acceptable, proceed
with the analyses of the source samples. If
they do not agree within 10 percent, then
determine the reason for the discrepancy,
and take corrective action before proceed-
ing.
d-Bws)
18-5
7.1.8 Emission Calculations. From the av-
erage calibration curve described In Section
7.1.5., select the value of C, that corresponds
to the peak area. Calculate the concentra-
tion C, In ppm, dry basis, of each organic In
the sample as follows:
c - csprT1Fr
LC " TT
where:
C.- Concentration of the organic from the
calibration curve, ppm.
P,«Reference pressure, the barometric
pressure or absolute sample loop pres-
sure recorded during calibration, mm
Hg.
TI-Sample loop temperature at the time of
sample analysis, 'K.
Fr-Relative response factor (if applicable,
see Section 6.4).
Pi-Barometric or absolute sample loop
pressure at time of sample analysis, mm
Hg.
T,-Reference temperature, the tempera-
ture of the sample loop recorded during
calibration,'K.
B.," Water vapor content of the bag sample
or stack gas, proportion by volume.
7.2 Direct Interface Sampling and Analy-
sis Procedure. The direct Interface proce-
dure can be used provided that the moisture
content of the gas does not interfere with
the analysis procedure, the physical require-
ments of the equipment can be met at the
site, and the source gas concentration Is low
enough that detector saturation Is not a
problem. Adhere to all safety requirements
with this method.
7.2.1 Apparatus.
7.2.1.1 Probe. Constructed of stainless
steel, Pyrex glass, or Teflon tubing as re-
quired by duct temperature, 6.4-mm OD, en-
larged at duct end to contain glass wool
plug. If necessary, heat the probe with heat-
Ins tape or a special heating unit capable of
maintaining duct temperature.
7.2.1.2 Sample Lines. 6.4-mm OD Teflon
lines, heat-traced to prevent condensation
of material.
7.2.1.3 Quick Connects. To connect
sample line to gas sampling valve on GC In-
strument and to pump unit used to with-
draw source gas. Use a quick connect or
equivalent on the cylinder or bag containing
calibration gas to allow connection of the
calibration gas to the gas sampling valve.
7.2.1.4 Thermocouple Readout Device.
Potentiometer or digital thermometer, to
measure source temperature and probe tem-
perature.
7.2.1.5 Heated Gas Sampling Valve. Of
two-position, six-port design, to allow
sample loop to be purged with source gas or
to direct source gas into the OC Instrument.
7.2.1.6 Needle Valve. To control gas sam-
pling rate from the source.
-------
Section No. 3.16.10
Date June 30, 1988
Page 8
o
7.2.1.7 Pump. Leakless Teflon-coated dia-
phragm-type pump or equivalent, capable of
at least 1 liter/minute sampling rate.
7.2.1.0 Flowmeter. Of suitable ranee to
measure sampling rate.
7.2.1.0 Charcoal Adsorber. To adsorb or-
ganic vapor collected from the source to
prevent exposure of personnel to source gas.
7.2.1.10 Gas Cylinders. Carrier gas
(helium or nitrogen), and oxygen and hy-
drogen for a flame lonization detector (FID)
If one la used.
7.2.1.11 Oas Chromatograph. Capable of
being moved into the field, with detector.
heated eta sampling valve, column required
to complete separation of desired compo-
nents, and option for temperature program-
ming.
7.2.1.12 Recorder/Integrator. To record
results.
7.2.2 Procedure. To obtain a sample, as-
semble the sampling system as shown in
Figure 18-12. Make sure all connections are
tight. Turn on the probe and sample line
heaters. As the temperature of the probe
and heated line approaches the source tem-
perature as Indicated on the thermocouple
readout device, control the heating to main-
tain a temperature of 0 to 3'C above the
eource temperature. While the probe and
heated line are being heated, disconnect the
sample line from the gas sampling valve,
and attach the line from the calibration gas
mixture. Flush the sample loop with cali-
bration gas and analyze a portion of that
tnsa. Record the results. After the calibra-
tion gas sample has been Hushed into the
OC Instrument, turn the gas sampling valve
to flush position, then reconnect the probe
sample line to the valve. Place the inlet of
the probe at the centroid of the duct, or at a
point no closer to the walls than 1 m, and
draw source gas into the probe, heated line,
and sample loop. After thorough flushing,
analyze the sample using the same condi-
tions as for the calibration gas mixture.
Repeat the analysis on an additional
sample. Measure the peak areas for the two
samples, and If they do not agree to within 8
percent of their mean value, analyze addi-
tional samples until two consecutive analy-
ses meet this criteria. Record the data.
After consistent results are obtained,
remove the probe from the source and ana-
lyze a second calibration gas mixture.
Record this calibration data and the other
required data on the data sheet shown in
Figure 18-11. deleting the dilution gas infor-
mation.
(Non: Take care to draw all samples, cali-
bration mixtures, and audits through the
sample loop at the same pressure.)
7.2.3 Determination of Stack Gas Mois-
ture Content. Use Method 4 to measure the
stack gas moisture content.
7.2.4 Quality Assurance. Same as Section
7.1.7. Introduce the audit gases in the
sample line immediately following the
probe.
7.2.5 Emission Calculations. Same as Sec-
tion 7.1.8.
7.3 Dilution Interface Sampling and
Analysis Procedure. Source samples that
contain a high concentration of organic ma-
terials may require dilution prior to analysis
to prevent saturating the OC detector. The
apparatus required for this direct interface
procedure is basically the same as that de-
scribed In the Section 7.2, except a dilution
system is added between the heated sample
line and the gas sampling valve. The appa-
ratus is arranged so that either a 10:1 or
100:1 dilution of the source gu can be di-
rected to the Chromatograph. A pump of
larger capacity is also required, and this
pump must be heated and placed in the
system between the sample line and the di-
lution apparatus:
7.3.1 Apparatus. The equipment required
in addition to that specified for the direct
Interface system is as follows:
7.3.1.1 Sample Pump. Leakless Tenon-
coated diaphragm-type that can withstand
being heated to 120'C and deliver 1.5 liters/
minute.
7.3.1.2 Dilution Pumps. Two Model A-160
Komhyr Teflon positive displacement type
delivering ISO cc/mlnute. or equivalent. As
an option, calibrated flowmeters can be used
In conjunction with Teflon-coated dia-
phragm pumps.
7.3.1.3 Valves. Two Teflon three-way
valves, suitable for connecting to 6.4-mm
OD Teflon tubing.
7.3.1.4 Flowmeters. Two. for measure-
ment of diluent gas, expected delivery flow
rate to be 1.3SO cc/min.
7.3.1.5 Diluent Oas with Cylinders and
Regulators. Oas can be nitrogen or clean
dry air, depending on the nature of the
source gases.
7.3.1.0 Heated Dox. Suitable for being
heated to 120'C. to contain the three
pumps, three-way valves, and associated
connections. The box should be equipped
with quick connect fittings to facUitate con-
nection of: (1) The heated sample line from
the probe, (2) the gas sampling valve. (3)
the calibration gas mixtures, and (4) diluent
gas lines. A schematic diagram of the com-
ponents and connections is shown in Figure
18-13.
(NOTE Care must be taken to leak check
the system prior to the dilutions so as not to
create a potentially explosive atmosphere.)
The heated box shown in Figure 10-13 la
designed to receive a heated line from the
probe. An optional design is to build a probe
unit that attaches directly to the heated
box. In this way, the heated box contains
the controls for the probe heaters, or. If the
box is placed against the duct being sam-
pled, it may be possible to eliminate the
probe heaters. In either case, a heated
Teflon line Is used to connect the heated
box to the gas sampling valve on the chro-
ma tograph.
O
o
-------
Section No. 3.16.10
Date June 30, 1988
Page 9
7.3.2 Procedure. Assemble the apparatus
by connecting the heated box, shown in
Figure 18-13. between the heated sample
line from the probe and the gas sampling
valve on the chromatosraph. Vent the
source gas from the ges sampling valve dl-
rectly to the charcoal filter, eliminating the
pump and rot&meter. Heat the sample
probe, sample line, and heated box. Insert
the probe and source thermocouple to the
centrold of the duct, or to a point no closer
to the walla than 1 m. Measure the source
temperature, and adjust all beating units to
a temperature 0 to 3'C above this tempera-
ture. It this temperature is above the safe
operating temperature of the Teflon compo-
nents, adjust the heating to maintain a tem-
perature high enough to prevent condensa-
tion of water and organic compounds.
Verify the operation of the dilution system
by analyzing a high concentration gas of
known composition through either the 10:1
or 100:1 dilution stages, as appropriate. (If
necessary, vary the flow of the diluent gas
to obtain other dilution ratios.) Determine
the concentration of the diluted calibration
gxs using the dilution factor and the cali-
bration curves prepared In the laboratory.
Record the pertinent data on the data sheet
shown in Figure 18-11. If the data on the di-
luted calibration era are not within 10 per-
cent of the expected values, determine
whether the chromatosreph or the dilution
system la in error, and correct it. Verify the
GC operation using a low concentration
standard by diverting the gas into the
isjnple loop, bypaeting the dilution Byctem.
If thesa analyzes are not within acceptable
limits, correct the dilution system to provide
the desired dilution factors. Make this cor-
rection by diluting a high-concentration
standard gas mixture to adjust the dilution
ratio as required.
Once the dilution system and QC oper-
ations are satisfactory, proceed with the
analysis of source gas. maintaining yje gaxat
dilution tattlnga as used for the standards.
Repeat the analyses until two consecutive
values do not vary by more than S percent
from their mean value are obtained.
Repeat the analysis of the calibration gat
mixtures to verify equipment operation.
Analyze the two field audit samples using
either the dilution system, or directly con-
nect to the gas campling valve as required.
Record all data and report the results to the
audit supervisor.
7.3.3 Determination of stack Gas Mois-
ture Content. Same ta Section 7.2.3.
7.3.4 Quality Assurance. Same as Section
7.2.4.
7.3.5 Emission Calculations. Same as Sec-
tion 7.2.5, with the dilution factor applied.
7.4 Adsorption Tube Procedure (Alterna-
tive Procedure). It is suggested that the
tester refer to the National Institute of Oc-
cupational Safety and Health (NIOSH)
method for the particular organica to be
sampled. The principal interferent will be
water vapor. If water vapor is present at
concentrations above 3 percent, silica gel
should be used In front of the charcoal.
Where more than one compound is present
in the emissions, then develop relative ad-
sorptive capacity information.
7.4.1 Additional Apparatus. In addition
to the equipment listed in the NIOSH
method for the particular orcanicKs) to bo
sampled, the following items (or equivalent)
are suggested.
7.4.1.1 Probe (Optional). BoroslUcate
glass or stainless steel, approximately 6-mm
ID, with a heating system if water conden-
sation is a problem, and a filter (either in-
stack or out-stack heated to stack tempera-
ture) to remove particulate matter. In most
instances, a plug of glass wool is a eatisfac-
tory filter.
7.4.1.2 Flexible Tubing. To connect probe
to adsorption tubes. Ues a material that ex-
hibits *T»<"i™*i sample adsorption.
7.4.1.3 Leakiest Sample Pump. Flow con-
trolled. constant1 rate pump, with a eet of
limiting (sonic) orifices to provide pumping
rates from approximately 10 to 100 cc/mln.
7.4.1.4 Bubble-Tube Flowmeter. Volume
accuracy within ± 1 percent, to calibrate
pump.
7.4.1.5 Stopwatch. To time »«pip""g and
pump rate calibration.
7.4.1.6 Adsorption Tubes. Similar to ones
specified by NIOSH. except the amounts of
adsorbent per primary/backup sections ore
eoo/aoo mg for charcoal tubas and KHO/2JO
ms for dlica eel tubes. As an alternative.
the tubes may contain a porous polymer ad-
sorbent such &a Tenax GC or XAD-2.
7.4.1.7 Barometer. Accurate to 5 mm Eg,
to measure atmospheric pressure during
sampling end pump calibration.
7.4.1.8 Rotemeter. 0 to 100 cc/mln, to
detect changes In flow rate during sampling.
7.4.2 Sampling and Analyria. It is sug-
gested that the tester follow the sampling
and analysis portion of the respective
NIOSH method section entitled "Proce-
dure." Calibrate the pump and limiting ori-
fice flow rate through adsorption tubes with
the bubble tube flowmeter before campling.
The sample system can be operated es a "re-
drculatiaff loop" for this operation. Record
the ambient temperature and barometric
pressure. Then, during sampling, use the ro-
tameter to verify that the pump and orifice
sampling rate remains constant.
Use a sample probe, if required, to obtain
the sample at the centroid of the duct, or at
a point no closer to the walls than 1 m. Min-
imize the length of flexible tubing between
the probe and adsorption tubes. Several ad-
sorption tubes can be connected in series, if
the extra adsorptive capacity is needed. Pro-
vide the gas sample to the sample system at
a pressure sufficient for the limiting orifice
to function as a sonic orifice. Record the
total time and sample flow rate (or the
number of pump strokes), the barometric
pressure, and ambient temperature. Obtain
a total sample volume commensurate with
the expected concentration^) of the volatile
organlc(s) present, and recommended
sample loading factors (weight sample per
weight adsorption media). Laboratory tests
prior to actual sampling may be necessary
to predetermine this volume. When more
than one organic is present In the emissions,
then develop relative adsorptive capacity in-
formation. If water vapor is present in the
sample at concentrations above 2 to 3 per-
cent, the adsorptive capacity may be severe-
-------
Section No. 3.16.10
Date June 30, 1988
Page 10
o
ly reduced. Operate the gas chromatograph
according to the manufacture's Instructions.
Alter establishing optimum conditions,
verify and document these conditions
during all operations. Analyze the audit
samples (see Section 7.4.4.3), then the emis-
sion samples. Repeat the analyst* of each
sample until the relative deviation of two
consecutive Injections does not exceed 5 per-
cent.
7.4.3 Standards and Calibration. The
standards can be prepared according to the
respective NIOSH method. Use a minimum
of three different standards; select the con-
centrations to bracket the expected average
sample concentration. Perform the calibra-
tion before and after each day's sample
analyses. Prepare the calibration curve by
using the least squares method.
7.4.4 Quality Assurance.
7.4.4.1 Determination of Desorptlon Effi-
ciency. During the testing program, deter-
mine the desorptlon efficiency In the ex-
pected sample concentration range for each
batch of adsorption media to be used. Use
an Internal standard. A minimum desorp-
tlon efficiency of 50 percent shall be ob-
tained. Repeat the desorptlon determina-
tion until the relative deviation of two con-
secutive determinations does not exceed 5
percent. Use the average desorption effi-
ciency of these two consecutive determina-
tions for the correction specified In Section
7.4.4.S. If the desorptlon efficiency of the
compound(s) of Interest Is questionable
under actual sampling conditions, use of the
Method of Standard Additions may be help-
ful to determine this value.
7.4.4.2 Determination of Sample Collec-
tion Efficiency. For the source samples, ana-
lyze the primary and backup portions of the
adsorption tubes separately. If the backup
portion exceeds 10 percent of the total.
amount (primary and backup), repeat the
sampling with a larger sampling portion.
7.4.4.3 Analysis Audit. Immediately
before the sample analyses, analyze the two
audits In accordance with Section 7.4.2. The
analysis audit shall agree with the audit
concentration within 10 percent.
7.4.4.4 Pump Leak Checks and Volume
Flow Rate Checks. Perform both of these
checks immediately after sampling with all
sampling train components in place. Per-
form all leak checks according to the manu-
facturer's Instructions, and record the re-
sults. Use the bubble-tube flowmeter to
measure the pump volume now rate with
the orifice used in the test sampling, and
the result. If It has changed by more than 5
but less than 20 percent, calculate an aver-
age flow rate for the test. If the flow rate
has changed by more than 20 percent, reca-
librate the pump and repeat the sampling.
7.4.4.5 Calculations. All calculations can
be performed according to the respective
NIOSH method. Correct all sample volumes
to standard conditions. If a sample dilution
system has been used, multiply the results
by the appropriate dilution ratio. Correct all
results by dividing by the desorptlon effi-
ciency (decimal value). Report results as
ppm by volume, dry basis.
7.5 Reporting of Results. At the comple-
tion of the field analysis portion of the
study, ensure that the data sheets shown In
Figure 18-11 have been completed. Summa-
rize this data on the data sheets shown In
•Figure 18-15.
8. Bibliopraphv
1, American Society for Testing and Mate-
rials. Ci Through C. Hydrocarbons in the
Atmosphere by Gas Chromatography.
ASTM D 2820-72. Part 23. Philadelphia, Pa.
23:950-988.1673.
2. Corazon, V. V. Methodology for Collect-
Ing and Analyzing Organic Air Pollutants.
UJ3. Environmental Protection Agency.
Publication No. EPA-600/2-79-042. Febru-
ary 1979.
3. Dravnieks, A.. B. K. Krotoszynskl. J.
Whltfleld, A. O'Donnell, and T. Burgwald.
Environmental Science and Technology.
SU2):1200-1222.1971.
4. Eggertsen. F. T., and F. M. Nelsen. Oaa
Chromatographic Analysis of Engine Ex-
haust and Atmosphere. Analytical Chemis-
try. J
-------
Section No. 3.16.10
Date June 30, 1988
Page 11
19. NIOSH Manual of Analytical Methods.
Volumes 1, 2, 3, 4, 5, 6, 7. U.S. Department
of Health and Human Services National In-
stitute for Occupational Safety and Health.
Center for Disease Control. 4676 Columbia
Parkway, Cincinnati. Ohio 45226. April
1977-August 1981. May be available from
the Superintendent of Documents. Govern-
ment Printing Office, Washington, DC
20402. Stock Number/Price: Volume 1—017-
033-00267-3/$13. Volume 2—017-033-00260-
6/411. Volume 3-017-033-00261-4/$l4,
Volume 4—017-033-00317-3/S7.25, Volume
5—017-033-00349-1/$10, Volume 6—017-033-
00360-6/J9, and Volume 7—017-033-00386-
5/$7. Prices subject to change. Foreign
orders add 25 percent.
20. Schuetzle. D.. T. J. Prater, and S. R.
Ruddell. Sampling and Analysis of Emis-
sions from Stationary Sources; I. Odor and
Total Hydrocarbons. Journal of the Air Pol-
lution Control Association. 25(9):925-932.
1975.
21. Snyder, A. D., P. N. Hodgson. M. A.
Kemmer and J. R. McKendree. Utility of
Solid Sorbents for Sampling Organic Emis-
sions from Stationary Sources. U.S. Envi-
ronmental Protection Agency. Research Tri-
angle Park. NC Publication No. EPA 600/2-
76-201. July 1976. 71 p.
22. Tentative Method for Continuous
Analysis of Total Hydrocarbons in the At-
mosphere. Intersociety Committee, Ameri-
can Public Health Association. Washington,
DC 1972. p. 184-186.
23. Zwerg, G., CRC Handbook of Chroma-
tography. Volumes I and II. Sherma. Joseph
(ed.). CRC Press. Cleveland. 1972.
-------
cf
Ccntscts
Process to be
Duct CP vsnt to ba i6=pled_
II. Process description
Rsw oatertal
Products
Cats
Pnsns
Operating cycla
Cheek: Bitch Continuous
fining of batch or cyclo
Cist tics to test
.Cyclic
Figure 18-1. Prsllninary survey data sheet.
Cespenanta to bo analysed gxpaetod eoneantration
Suggoatad chxoaatogxaphie
Coltm flew rato nl/nin Hoad prosoora t
Colum toaparatnroi
laothamal *C
Progra==3d fren *C to _ 'C at _
Injaction port/aaopla loop terrporaturo _ ^*C
Datactor temperature *C
Datactor flow rataai Hydrogen _____nl/nin.
haad praaauro
Air/Osygen nl/aia,
hoad pr a a aura
Chart apaed _________ inchea/ninuta
Cerpoond datat
Conpound Ratantien tlna Attanuatien
39
Hg
rsa Hg
TJ o C/)
£3 p CD
oq rf o
CD CD ct
O
P
fu
7igara 10-2. Chrcaatographic eonditiono data ohaat.
0) 2!
O
CO •
o
- OJ
CX>-
OO l
O
o
o
-------
Preparation of Standards In Tedlar Bags
and Calibration Curve
Standards
Mixture" Mixture Mixture
Standards Preparation Data: II »2 13
Organic:
Bag timber or identification
Dry gas ester calibration factor
Final neter reading (liters)
Initial eater reading (liters)
Hetered voluse (liters)
Average eeter teaperature (*K)
Average ester pressure, gauge (rai Hg)
Average ataospherlc pressure (ea Hg)
Average ester pressure, absolute (ea Hg)
Syringe temperature (*X)
(Section 6.2.2.1)
Syringe pressure, absolute (ea Hg)
(Section 6.2.2.1)
YoTuae of gas In syringe (nl)
(Section 6.2.2.1)
Density of liquid organic (g/nl)
(Section 6.2.2.2)
Voluae of liquid In syringe ({!)
(Section 6.2.2.2)
GC Operating Conditions:
Stsple loop voline (•!)
Saaple loop tesperature (*C)
Carrier gas flow rate (tri/nln)
Coluan tcsperature
Initial CO
Rate change PC/Bin)
Final CCJ
Organic Peak Identification and
Calculated'Concentrations:
Injection ties (24-hr clock)
Distance to paak (ca)
Chart speed (ca/nln)
Organic retention tiee (nln)
Attenuation factor
Peak height (n)
Peak area («2)
Peak area x attenuation factor
Calculated concentration (ppa)
(Equation 18-3 or 18-4)
Plot peak area x attenuation factor against calculated concentration
to obtain calibration curve.
rtgure 18-3. Standards prepared In Tedlar bags
and calibration curve.
Flcvatter Calibration
Flcwseter nusber or Identification
Flowsteter type
Calibration device (x):Bubble raeter
Readings at laboratory conditions:
Laboratory tesperature (Ti4(,)
Spl roaster
Laboratory baresttrie pressure (Pi*b'
Flow data:
Flowseter
Wet testracier
Hg
Calibration device
reading ~| tesp.
(as narked) I (*X)
i pressure
l(absolute)
Titse I I ~
(nln) Igas voltaa* jflow rate"
T I
I I
a » Voluae of gas neasured by calibration device, corrected to standard
conditions (liters).
b * Calibration device gas volcee/tln*.
Plot f! wester reading against flow rate (standard conditions), and draw a
seooth curve. If tns fletflMter being calibrated 1s • rottatter or otter
How device that Is viscosity dspenitnt, It wy be necessary to $tfier«tt a
•feally' of calibration curves that cover the operating pressure end
tcs?eraturt ranges of th» floustter.
tla following technique should be verified before application. It nay
bt possible to calculate flow rat* readings for rotczcters at standard
conditions Qjtd ts follows:
T1ab\ »«
b\
V
Flew rata
(lefearatcry cenditions)
Ret* rat«
(standard conditions)
Figure 18-4. Flotewtw calibration.
TJ a to
CD Co (D
oq ci- o
to a> ct-
H-
I-1 «-4 o
UJ C O
ro s:
o
uo •
o
VD a\
<»•
00 H'
o
-------
Section No. 3.16.10
Date June 30, 1988
Page 14
o
CAS
CYUHMIl
DILUENT
CAS
cnnron?
h«^
\
CALlDRATTDROTAf.'ITinS
nim ROB CONTROL
VAIVES
Flpjr» 1S-5.
calfbritfra
o
men
co:;cniTnATio:i
DASTE
mm—
3—f,lEtJl£ VALVES
J
PRESSURE
ion
om
GAS
DIUmiTAIR
DIUKHTAin
PURESUGSTAtlCEOn
MIXTURE
Figure 10-6. Tttostigs dllu.tlon epparatus.
O
-------
Section No. 3.16.10
Date June 30, 1988
Page 15
Preparation of Standards by Dilution of Cylinder Standard
Cylinder standard: Organic Certified concentration
Standards Preparation Data: Date
ppa
Stage 1 Mixture 1 Mixture 2 Mixture 3
Standard gas flewmeter reading
Diluent gas flowaeter reading . _____
Laboratory teaperature CK)
Baresstrlc pressure (o Hg)
Flotottter gage pressure (IBS Hg)
Flow rate cylinder gas at
standard conditions (al/nln)
Flow rate diluent gas at
standard conditions (nl/raln)
Calculated concentration (ppa)
Stage 2 (If used)
Standard gas flowaeter reading
Diluent gas flowaater reading
Flow rate stage 1 gas at
standard conditions (al/nln)
Flow rate diluent gas at
standard conditions (ml/Bin)
Calculated concentration (ppn)
GC Operating Conditions:
Sample loop voluae (nl)
Saaple loop temperature CO
Carrier gas flow rate (nl/otn)
Column tenperature:
Initial CO
Prograa rate CC/nln) "~~~~
••: F1nal (
-------
Section No. 3-16.10
Date June 30, 1988
Page 16
COILING
WATER
OATH
SYRINGE
SEPTUM
4- MIDGET
IMPIKGER
HOTPLATE
NITROGEN
CniNDER
o
Figure 18-0. Apparatus for preparation cf liquid eatcrlali.
VZtJT
TIRO?!
STACK
HAU.
FRTIR fl
C31ASS HJOll 11
nrvinst
O-JIYPE
PITOTTUEE
RIGID UAKRJ007 (XWTAI
O
Figure 10-9. Integrated bag strpHng train.
-------
PrcSa
8* Tefla
WstftClta?
Section No. 3.16.1C
Date June 30, 1988
Page 17
Mr Tlfht Steal Crta „
«
p* oHtpa j
/--..
*:
:
' 0 1
\ ' :
« . •
% :
i
/
«._v
Hf&x
EtatgateJ Steal
Crca
18-9*. Exploslca risk CM (telling esthsd.
P1«J»t_
S1ta_
Oats
Simple g
Source tcrpsraturo (*C)
Testtrrotca Hg)_
tc=?«ratara (*C)
Sample flcy rate(eppr.)
Dag mnbsr
Start tlcz
Finish tlsa
Figure 18*10. Field staple data sheet - Tedlar
bag collection catted.
-------
Section No. 3.16.10
Date June 30, 1988
Page 18
o
Plant..
Location
Caneral 4nfornaU.cn
Source t*Ep*raturo (*C)
Probe tarperatture CO
AsMent tezporature (*C)
Atsoiphtrie praaaure (ea)
Coturce preaaero (*Cg)
Absolute eooreft prtaaura p««4 tea/nin)
Dilation gas flow rate (al/ein)
Dilution Ca« nsad (tyebol)
Dilution ratio
O
HCJTV 16-11. mid tnalyslt data stem.
a. ?ield «aalynia Data - Calibration Cue
Can ea. Tiia
Area Attenuation A x A Factor Cene.
torn Co. tlea
Area Attanaatien A »e A Taeter Cone, (ppn)
Ban Bo. tlea
Cenponanta Area Atttnuatien A « A Factor Cene. (ppn>
O
Flpirt 1>-11 (tcntlr.uad). Flald analysis dau th«tu.
-------
Section No. 3-16.10
Date June 30, 1988
Page 19
TC
aietai
- II
re Ktton
OS
DiKSCUCt
1
hlRCSSCS
18»J2. Direct Interfeca
tystcn.
Vent to Charcoal Adsorbers
Heated Line
Frosa Probe
' »*c
Quick
Connect
i
x \
-fr
t3
Source
Gas Pirrp
, i
iyj
cc/MIr
— (
1.5 L/M1n "
i
10:1
100:1
^^S
rW
£1
)
_f
3-Hay Valves
In 100:1
Position '
V -**
Quick Connects
To Gas SeKple
Valve
r— f-) 150 cc/MIn
)
Check Valve I
i Quick Connects I
> For Calibration L
i
H cssaters
(On Outside
Of eor.)
Row Rate Of
1350 cc/Ntn
Heated Box at 1ZO°C Or Source Temperature
Figure 18-13. Schematic dlagraa of the heated box required
for dilution of sarple gas.
-------
Section No. 3.16.10/*—\
Date June 30, 1988 ( )
Page 20 V-X
QACTOU* Oioinc BAxrwra jure AHALTIU
GnrsLirT
floWOV AoUTM flOCfM
Olcswod with Initial* or number u
•pproprUte) ' *~ <
LO*a*nl
Coorot
tsaotntttn
L rronrm dtu:
B. Onb aaete utlTMd" <"c>-
" ' ~ ——— uoscnton
CO
Ateiotpbcrte
t t^aontorT rmKtrmrton data: Coaret
X OtUbntiao curm pr»-
puvd—_^ O _
Number of eompo- (mlAami
tuntg, O ~
CimpUloop
conpootot (1 rv- U8p
B. Aattt »«iTjil«i (opttoo- Csmpl*
Amlnl> eonplM- UaM(M-lsr
Cotasn
t49ffiC
tare
O
«. ete»U>» preewtanK rCX.
XIKttMMt Precna
BM*UI«I*____ a a rsurc/
a o
CdstJaa tnUrtJ6t_ Q O FtnttlCC}..
B. ItmalMr of HB9>« eat- O»m«rf«a
tn flovnl*
4. rwkl u^trttz Unl/BtoX
X TBttl bjdi'oeiiton DtUctar
KambCT of ersrapo-
1S*14. fifnpllny tnd uulysli
Ftrformod by Ulm»lnr«l-
auxou* Otcunc Rtvnna uro Axu.Tti* __
DATA DU« -
flow rmtc
pUnt
pm PUure 18-14. o«Bipaaj «ad (nairtli sheet.
O
-------
Section No. 3.16.10
Date June 30, 1988
Page 21
APPENDIX C—QUALITY ASSURANCE
PROCEDURES*
Procedure 1—Determination of Adequate
ChromatosrrapMc Peak Resolution
In this method of dealing with resolution.
the extent to which one chromatographic
peak overlaps another la determined.
For convenience, consider the ranee of the
elution curve of each compound as running
from -2
-------
Section No. 3.16.10
Date June 30, 1988
Page 22
o
C *fc_«_ k_
b-JOj b-to, b*2of
-07 -07-
Th* following calculation ttept art r*qu1red:*
1. 2o$ « t./^ In 2
2. oe • t£
3. x, « (b
4. x,
S. fl(x,)»-i
P-x*\
"*" /dx
8. A0 « IflAe/A,
9. Pircentag* overlap « Afl x 100 ,
A •
AC «
tc «
C(xi) *
Q(*«) *
Ifl «
Ar«a of th* i«tplt p«tk of Inttrtft daUralntd by tltctronlc inte-
gration or by th* forcula A( » h t..
Arta of th* contnlnant ptik, daunltttd In th* tax aanntr at A(.
Olitane* on th* chrcutographfc chart that ttparatat the uxlu of
thj two p«aki.
Pttk h*1eht of th* ttspU coxpound of 1nt*r*>t, M*tur*d froa th*
«v«r«j» vilu* of th* paialtn* to th» BUtnui of th* curv*.
Width of ««cpU p*ak of InUrcit at 1/2 p*ak h**xJ«r preesur*. p«_
ij/bdux
D.ppm.
lOKCone.
It^l (JUflC.
o
Part S.—To be filled out by audit mpetvl-
sor. - . .
I. Process sampled
2. Audit location-
3. Name of individual audlt-
4. Audit date
5. Audit results:
b. Cy£ndv pnMturt before audit, pet
c. Cy&ndor pf*»eur* after eurfit p*^
fl* fc^ecton tC2* Avere^e
*. Actual audK eonoerrtratioa Ppm (Part A,
«•}...
f. Aucflecctncy:1
ImrCmw. Cyimtif
Hijfi Cwc, CV"^1**
Percent' aceureey -
Ueejind Conc-Actut* Cone.
AdudConc.
Q pfi^r"!
-------
Section No. 3.16.11
Date June 30, 1988
Page 1
11.0 REFERENCES
1.
2.
3.
4.
Method 18 - Measurement of Gaseous Organic Compound - Emissions by Gas
Chromatography. Federal Register, Volume 48, No. 202, October 18, 1983 ,
page
5.
6.
7.
8.
9.
10.
11.
12.
Amendments to Method 18. Federal Register, Volume 49, No. 105, May 30,
1984, page 22608.
Miscellaneous Clarifications and Addition of Concentration Equations to
Method 18. Federal Register. Volume 52, No. 33, February 19, 198?, page
5105.
Stability of Parts-Per-Million Organic Cylinder Gases and Results of
Source Test Analysis Audits, Status Report #8. U. S. Environmental
Protection Agency Publication No. EPA-600/2-86-117, January 198?. Also
available from NTIS as Publication No. PB 8y-l4l46l.
Traceability Protocol for Establishing True Concentration of Gases Used
for Calibration and Audits of Continuous Source Emission Monitors
(Protocol No. 1). Section 3-0.4, Quality Assurance Handbook, Volume
III, Stationary Source Specific Methods, U. S. Environmental Protection
Agency Publication No. EPA-600/4-?7-027b, June 15, 1978.
Methanol, Method 2000. NIOSH Manual of Analytical Methods, Volume 2,
Third Edition, U. S. Department of Health and Human Services, February
1984.
Alcohols I, Method 1400. NIOSH Manual of Analytical Methods, Volume 1,
Third Edition, U. S. Department of Health and Human Services, February
1984.
Alcohols II, Method 1401. NIOSH Manual of Analytical Methods, Volume 1,
Third Edition, U. S. Department of Health and Human Services, February
1984.
Hydrocarbons. BP 36 - 126° C, Method 1500. NIOSH Manual of Analytical
Methods, Volume 2, Third Edition, U. S. Department of Health and Human
Services , February 1984 .
Development of Methods for Sampling 1,3-Butadiene. Interim Report
prepared under U. S. "Environmental Protection Agency Contract Number
68-02-3993, March 1987-
Hexachlorocyclopentadiene, Method 2518. NIOSH Manual of Analytical
Methods, Volume 2, Third Edition, U. S. Department of Health and Human
Services, February 1984.
Method 110 - Determination of Benzene from Stationary Sources, Proposed
Rule. Federal Register. Volume 45, No. 77, April 18, 1980, page 26677.
-------
Section No. 3.16.11
Date June 30, 1988 >«.
Page 2 f j
13. Hydrocarbons, Aromatic, Method 1501. NIOSH Manual of Analytical Methods,
Volume 2, Third Edition, U. S. Department of Health and Human Services,
February 1984.
14. Naphthylamines, MetftG>a£li264. NIOSH Manual of Analytical Methods, Volume
4, Second Edition, U. S. Department of Health and Human Services, August
1978.
15. Ketones I, Method 1300. NIOSH Manual of Analytical Methods, Volume 2,
Third Edition, U. S. Department of Health and Human Services, February
1984.
16. 2-Butanone, Method 2500. NIOSH Manual of Analytical Methods, Volume 1,
Third Edition, U. S. Department of Health and Human Services, February
17- Ethylene Oxide, Method 1607. NIOSH Manual of Analytical Methods, Volume
1, Third Edition, U. S. Department of Health and Human Services, February
1984.
18. Propylene Oxide, Method 1612. NIOSH Manual of Analytical Methods. Volume
2, Third Edition, U. S. Department of Health and Human Services, February
1984.
19. Hydrocarbons, Halogenated, Method 1003. NIOSH Manual of Analytical
Methods, Volume 2, Third Edition, U. S. Department of Health and Human
Services, February 1984.
20. Ethylene Dibromide, Method 1008. NIOSH Manual of Analytical Methods,
Volume 1, Third Edition, U. S. Department of Health and Human Services,
February 1984.
21. Method 23 - Determination of Halogenated Organics from Stationary Sources
(proposed method). Federal Register. Volume 45. No. 114, June 11, 1980,
page 39766.
22. . 1,2-Dichloropropane, Method 1013. NIOSH Manual of Analytical Methods,
Volume 1, Third Edition, U. S. Department of Health and Human Services,
February 1984.
23. Development of Methods for Sampling Chloroforn and Carbon Tetrachloride.
Interim Report prepared for U. S.. Environmental Protection Agency under
EPA Contract Number 68-02-3993, November 1986.
24. Dichlorodifluoromethane, Method 111. NIOSH Manual of Analytical Methods,
Volume 2, Second Edition, U. S. Department of Health and Human Services,
April 1977.
O
O
-------
Section No. 3.16.11
Date June 30, 1988
Page 3
25. Methyl Bromide, Method 2520. NIOSH Manual of Analytical Methods, Volume
2, Third Edition, U. S. Department of Health and Human Services, February
1984.
26. Methyl Chloride, Method 99. NIOSH Manual of Analytical Methods, Volume
4, Second Edition, U. S. Department of Health and Human Services, August
1978.
27. Butler, F. E., E. A. Coppedge, J. C. Suggs, J. E. Knoll, M. R. Midgett,
A. L. Sykes, M. W. Hartiaan, and J. L. Steger. Development of a Method
for Determination of Methylene Chloride Emissions at Stationary Sources.
Paper for presentation at 80th Annual Meeting of Air Pollution Control
Association, New York, NY, June 1987.
28. Vinylidene Chloride, Method 266. NIOSH Manual of Analytical Methods,
Volume 4, Second Edition, U. S. Department of Health and Human Services,
August 1978.
29. Ethyl Chloride, Method 2519. NIOSH Manual of Analytical Methods, Volume
1, Third Edition, U. S. Department of Health and Human Services, February
1984.
30. Method 106 - Determination of Vinyl Chloride from Stationary Sources.
Federal Register, Volume 47, No. 173, September 7, 1982, page 39168.
31. Knoll, J. E., M. A. Smith, and M. R. Midgett. Evaluation of Emission
Test Methods for Halogenated Hydrocarbons, Volume II, U. S. Environmental
Protection Agency Publication No. EPA-600/4- 80-003, January 19&0.
32. Methylene Chloride, Method 1005. NIOSH Manual of Analytical Methods,
Volume 2, Third Edition, U. S. Department of Health and Human Services,
February 1984.
33. Tetrachloroethylene, Method 335. NIOSH Manual of Analytical Methods,
Volume 3» Second Edition, U. S. Department of Health and Human Services,
April 1977-
34. Trichloroethylene, Method 336. NIOSH Manual of Analytical Methods,
Volume 3t Second Edition, U. S. Department of Health and Human Services,
April 1977.
35. -1,1,2-Trichlorotrifluoroethane. Method 129. NIOSH Manual of Analytical
Methods, Volume 2, Second Edition, U. S. Department of Health and Human
Services, April 1977- .
36. Vinyl Chloride, Method 1007. NIOSH Manual of Analytical Methods, Volume
2, Third Edition, U. S. Department of Health and Human Services, February
1984.
37- Mann, J. B., J. J. Freal, H. F. Enos, and J. X. Danauskas. Development
and Application of Methodology for Determining 1,2 Dibromo-3-
Chloropropane (DBCP) in Ambient Air. Journal of Environmental Science
and Health, B15(5), 519-528 (1980).
-------
Section No. 3.16.1
Date June 30. 1988
Page 4
38. VOC Sampling and Analysis Workshop, Volume III. U. S. Environmental
Protection Agency Publication No. EPA-340/1-84-001C, September 1984.
39. Knoll, J. E., M. A. Smith, and M. R. Midgett. Evaluation of Emission Test
Methods for Halogenated Hydrocarbons, Volume I. U. S. Environmental
Protection Agency Publication No. EPA-600/4-79~025, March 1979.
40. Binetti, R. et al. Headspace Gas Chromatographic Detection of Ethylene
Oxide in Air. Chromatographia, Vol. 21, December 1986.
111. Butadiene, Method 591. NIOSH Manual of Analytical Methods, Volume 2,
Second Edition, U. S. Department of Health and Human Services, April 1977.
42. Knoll, J. E. Estimation of the Limit of Detection in Chromatography.
Journal of Chromatographic Science, Vol. 23, September 1985.
43. Procedure 1 - Determination of Adequate Chromatographic Peak Resolution.
Code of Federal Regulations, Title 40. Part 6l. Appendix C, July 1, 1987-
44. Method 625 - Base/Neutrals and Acids. Code of Federal Regulations, Title
40. Part 136. Appendix A, July 1, 1987.
45. C. through C_ Hydrocarbons in the Atmosphere by Gas Chromatography. ASTM I/ J
2820-72, Part 23. American Society for Testing and Materials, ^—/
Philadelphia, PA, 23:950-958, 1973.
46. Corazon, V. V. Methodology for Collecting and Analyzing Organic Air
Pollutants. U. S. Environmental Protection Agency Publication No.
EPA-600/2-79-042, February 1979.
47. Dravnieks, A., B. K. Krotoszynski, J. Whitfield, A. O'Donnell, and T.
Burgwald. Environmental Science and Technology, 5(12):1200-1222, 1971.
48. Eggertsen, F. T., and F. M. Nelson. Gas Chromatographic Analysis of Engine
Exhaust and Atmosphere. Analytical Chemistry, 30(6): 1040-1043, 1958.
49. Feairheller, W. R., P. J. Marn, D. H. Harris, and D. L. Harris. Te<3inical
Manual for Process Sampling Strategies for Organic Materials, U. S.
Environmental Protection Agency, Publication No. EPA 600/2-76-122. April
1976.
50. FR, 39 FR 9319-9323, 1974.
51. FR. 39 FR 32857-32860. 1974.
52. FR, 41 FR 23069-23072 and 23076-23090, 1976.
53. FR, 41 FR 46569-46571, 1976.
54. FR, 42 FR 41771-41776, 1977.
-------
Section No. 3.16.11
Date June 30, 1988
Page 5
55. Fishbein, L. Chromatography of Environmental Hazards, Volume II. Elsevier
Scientific Publishing Company, New York, New York, 1973-
56. Hamersma, J. W., S. L. Reynolds, and R. F. Maddalone. EPA/IERL Procedures
Manual: Level 1 Environmental Assessment, U. S. Environmental Protection
Agency Publication No. EPA 600/276/l60a, June 1976.
57. Harris, J. C., M. J. Hayes, P. L. Levins, and D. B. Lindsay. EPA/IERL
Procedures for Level 2 Sampling and Analysis of Organic Materials. U. S.
Environmental Protection Agency Publication No. EPA 600/7-79-033, February
1979-
58. Harris, W. E., H. W. Habgood. Programmed Temperature Gas Chromatography.
John Wiley & Sons, Inc. New York, 1966.
59• Methods of Air Sampling and Analysis. Intersociety Committee, American
Health Association, Washington, D. C., 1972.
60. Jones, P. W., R. D. Grammer, P. E. Strup, and T. B. Stanford.
Environmental Science and Technology, 10:806-810, 1976.
61. McNair Han Bunelli, E. J. Basic Gas Chromatography. Consolidated Printers,
Berkeley, 1969.
62. Nelson, G. 0. Controlled Test Atmospheres, Principles and Techniques. Ann
Arbor, Ann Arbor Science Publishers, 1971.
63. Schuetzle, D., T. J. Prater, and S. R. Ruddell. Sampling and Analysis of
Emissions from Stationary Sources; I. Odor and Total Hydrocarbons.
Journal of the Air Pollution Control Association, 1975-
64. Snyder, A. D., F. N. Hodgson, M. A. Kemmer, and J. R. HcKendree. Utility
of Solid Sorbents for Sampling Organic Emissions from Stationary Sources.
U. S. Environmental Protection Agency Publication No. EPA 600/2-76-201,
July 1976.
65. Tentative Method for Continuous Analysis of Total Hydrocarbons in the
Atmosphere. Intersociety Committee, American Public Health Association,
Washington, D.C., 1972.
66. Zwerg, G. CRC Handbook of Chromatography, Volumes I and II. CRC Press,
Cleveland, 1972.
-------
o
o
o
-------
Section No. 3.16.12
Date June 30, 1988
Page 1
12.0 DATA FORMS
Blank data forms are provided on the following pages for the convenience of
the Handbook user. Each blank form has the custoaary descriptive title centered at
the top of the page. However, the section-page documentation in the top right-hand
corner of each page has been replaced with a number in the lower right-hand corner
that will enable the user to identify and refer to a similar filled-in fora in a
text section. For example, form M18-2.5 indicates that the form is Figure 2.5 in
Section $.18.2 of the Method 18 section. Future revisions of these forms, if any.
can be documented by 2.5a, 2.5b, etc. Nineteen of the blank forms listed below are
included in this section. Six have been left blank in the text as shown following
the form number.
Form
Title
Flowmeter Calibration Data Form
(English and metric units)
Critical Orifice Calibration Data Form
(English and metric units)
Dynamic Dilution Data Form
Static Dilution Data Form
Thermometer Calibration Forn
Preliminary Survey Data Sheet
Preliminary Survey Preparations
Pretest Sampling Checks
Pretest Preparations
Field Sampling Data Form for Container Sampling
Field Sampling Data Form for Direct Interface Sampling
Field Sampling Data Form for Adsorption Tube Sampling
On-site Measurements Checklist
Data Form for Analysis of Method 18 Samples
Calibration Standard Preparation Data Form for Diluted
Gas Cylinders
Calibration Data Form for Preparation of Standards in
Tedlar Bags by Gas and Liquid Injection
-------
Section No. 3.16.12
Date June 30, 1988
Page 2
o
5.8 Data Form for Development of Response and Relative
Retention Factors
5.9 Data Form for Preparation of Liquid Standards and
Desorption Efficiency Samples for Adsorption Tube Analysis
5.10 (Text) Postsampling Operations Checklist
6.1 Calculation Form for GO Analysis by Gas Injection
6.2 Calculation Form for GC Analysis by Liquid Injection
8.1 Field Audit Report Form
8.2 Method 18 Checklist to be Used by Auditors
O
o
-------
FLOWMETER CALIBRATION DATA FO
glish units)
Date
Calibrated by
Meter system no.
Barometric pressure, Pm =
Type of primary meter: wet test
Type of flowmeter calibrated: rotameter
in. Hg Ambient temperature
Primary meter no.
op
, dry gas
, or bubble meter
, dry gas meter
or mass flowtneter
Primary meter readings
Initial
reading
(vpl),»
ft3
Final
reading
(V),'
ft3
Initial
temp,°F
(tpi)
op
Final
temp,°F
(tpr>
oF
Press
drop
(Dp)<
in.
H20
Flowmeter readings
Initial
reading
-------
FLOWMETER CALIBRATION DATA FORM (metric units)
Date
Calibrated by
Meter systen no.
Primary deter no.
°C
Barometric pressure, Pn * mn Hg Ambient temperature
Type of primary meter: wet test , dry gas , or bubble meter
Type of flowmeter calibrated: rotaineter
, dry gas meter
, or mass flowmeter
Primary meter readings
Initial
reading
(vpi).a
m3
Final
reading
(vpf),a
D3
Initial
temp,°F
('PI*
°C
Final
temp,°F
(tpf)
°C
Pres
drop
(V
mm
^0
Flowmeter readings
Initial
reading
(V,i).b
Ej3 or
ia- /nin
Final
reading
(V"
n3 or
ia3/min
Initial
temp
(t.i)
°C
Final
temp
°C
Press
drop
(Ds).c
mm
H20
Time
min
(9)td
min
Calibration
factors
(Yt).e
(Y)
tt Volume passing through the neter using the initial and final readings and requires a minimum of at
least five revolutions of the meter.
b Volume passing through the meter using the initial and final readings or the indicated flow rate
using the initial and final flow rate setting.
0 Pressure drop through the meter used to calculate the neter pressure.
d The time it takes to complete the calibration run.
e With Y defined as the average ratio of volumes for the primary meter compared to the flowmeter
calibrated, Yj = Y + 0.03Y for the calibration and Yj = Y + 0.05Y for the posttest checks; thus,
For calibration of the dry gas neter:
- Vpiint.i * fc.f)/2 *'273°K][P. * {Dp/13.6)] _ YI
273"K][P.
For calibration of the rotaneter and nass flowaeter:
fc.f)/2 * 273°K][PB
(Eq. 2-6)
(Dp/13.6)]
o
tpf)/2 + 2738K][Pa
{Eq. 2-7), Y -
(Eq. 2-8)
o
Quality Assurance Handbook H18
-O
-------
CRITICAL ORIFICE CALIBRATION DATA FORM (English units)
Date
Calibrated by
Meter system no.
Primary meter no.
op
Barometric pressure, Pa = in. Hg Ambient temperature
Type of primary meter: wet test , dry gas , or bubble meter
Type of critical orifice: capillary glass
needle or tubing
, or adjustable
Primary meter readings
Initial
reading
(Vpi)/
ft3
Final
reading
(vpf),a
ft3
Initial
temp,°F
-------
CRITICAL ORIFICE CALIBRATION DATA FORM (metric units)
Date
Calibrated by
Meter system no.
Barometric pressure, PB =
Type of primary meter: wet test
Type of critical orifice: capillary glass
Primary neter no.
°C
mm Hg Ambient temperature
, dry gas , or bubble meter
needle or tubing
, or adjustable
Primary meter readings
Initial
reading
mm
H20
Critical orifice readings
Initial
setting
b
L or
L/min
Final
setting
b
L or
L/min
Press
drop
c
mm
Hg
Time
rain
(9),"
min
Calculated
flow rate
W(.td)]'
L/min
Calibration
factor'
(K'J
(K')
* Volume passirig through the meter using the initial and final readings and requires a minimum of at
least five revolutions of the meter.
b Volume passing through the orifice using the initial and final readings or the indicated flow rate
using the initial and final flow rate setting (for variable setting orifice only).
c Pressure drop through the meter used to calculate the meter pressure.
d The time it takes to complete the calibration run.
e With K' defined as the average orifice calibration factor based on the volumes of the primary test
meter, K't = K1.^ 0.03K1 for the calibration and K'j * K1 + 0.05K1 for the posttest checks; thus,
Flow rate of the primary meter at standard conditions:
0.3858(Vpf - Vpl)(P0 + Dp/13.6)
td)
tf)/2 * 2?3°C]
(Eq. 2-13), Q(ltd) =
pf
0
(Eq. 2-
For determination of the K1 for the critical orifice:
(Eq. 2-15), &
K'
K'
(Eq. 2-16)
• ba r
e
o
o
Quality Assurance Handbook Ml2-«2.2B
-------
DYNAMIC CALIBRATION DATA FORM
Date
Source flovnaeter number
Stage 1 flowoeter number
Stage 2 flovnaeter number
Barometric press mm (in.) Hg
Organic compound
Calibrated by
Date source meter calibrated _
Date stage 1 meter calibrated
Date stage 2 meter calibrated
Heated box temperature
Leak check for total system
Certified concentration
ppmv(X) Date of calibration curve
STAGE 1
Emission gas flowmeter reading, ml/min (q,.,).
Diluent gas flovnaeter reading, ml/min
Dilution ratio
Injection time, 24h
Distance to peak, cm
Chart speed, cm/min
Retention time, min
Attenuation factor
Peak area or units
Peak area X attenuation factor
Measured concentration,* ppmv
Calculated concentration," ppmv (C, )
Percent difference,6 % ,
RUN 1
RUN 2
STAGE 2 (if applicable)
Emission gas flovnaeter reading, ml/min
Diluent gas flowsaeter reading, ml/min
Dilution ratio
Injection tins, 24h
Distance to peak, cm
Chart speed, cm/min
Retention time, min
Attenuation factor
Peak area or units
Peak area X attenuation factor
Measured concentration,' ppmv
Calculated concentration,4 ppmv
Percent difference,6 %
RUN 1
RUN2
RUN3
v * See Figure 5-1 for calculation.
106 x (X x qc )
c Percent Difference
3 Calculated concentration for single stage
Calculated Concentration- - Measured Concentration
Measured Concentration
x 100J5
C. = 106 x X
= Calculated cone, for two stage
Quality Assurance Handbook M18-2.5
-------
Date
Source flowneter number
Dry gas meter number
Ambient temperature
Barometric press _
Organic compound
Certified concen,
STATIC DILUTION DATA FORM
Calibrated by
o
°C (°F)
mm (in.) Hg
(X)
ppmv
Date source meter calibrated _
Date dry gas meter calibrated
Dry gas meter calib factor (Y)
Leak check for total system _
Vacuum during leak check _ _
Date of calibration curve
Initial dry gas meter reading, L (ft3)
Final dry gas meter reading, L (ft3)
Volume of diluent gas metered, L (ft3)
Gas metered X calibration factor (Y),{V2}
Flowmeter sampling rate, ml/min (cfm)
Sampling time, nin
Sampling rate X sample tine , L ( ft3 ) , { Vt }
Dilution ratio
Injection time, 24h
Distance to peak, en
Chart speed, en/rain
Retention time, nin
Attenuation factor •
Peak area or units
Peak area X attenuation factor
Measured concentration,* ppmv
Calculated concentration,11 pprav, {Cs}
Percent difference,* %
RUN 1
RUN 2
* See Figure 5»i for calculations.
b Calculated concentration (C,) **
V2)
ppmv
c Percent difference, #d
Measured concent - Calculated concent
Measured concentration
X 100
The percent difference must be less than 10 % absolute.
Quality Assurance Handbook H18-2.6
O
-------
THERMOMETER CALIBRATION FORM
Date
Reference
thermometer
type
Calibi
thermc
type
•ated
>meter
use
no.
Ambler
refer*
it temper
calibrb
Meaaurec
*ature
differc
[ values
Bo:
refer*
i
iling wal
calibrb
;er
differc
Calibrator's
initials
Temperature reading of the reference thermometer in °C or °F.
Temperature reading of the thermometer being calibrated in °C or °F.
Difference between the reference thermometer and the calibrated thermometer. This difference must
be less than 3°C (5.4°F) for than initial calibration and 6°C (10.4°F) for the calibration check.
Quality Assurance Handbook M18-2.7
-------
FIELD SAMPLING DATA FORM FOR CONTAINER SAMPLING
Plant
City
Operator
Date
Flowneter calib.(Y)
Container type: bag
Run number
Stack dia, mm (in.)
Sample box number
syringe
canister
Container volume,
Container number
Average ( P)
Initial flowmeter setting
Average stack temp
Dilution system: (dynamic)
emission flowsetting
diluent flowsetting
Dilution system: (static)
liters emission flowsetting
Final leak check
m3/min (cfra)
mm (in.) IljO Vacuum during leak check
Pitot tube (Cp)
Static press mm (in.) HjO Barometric press
°C (°F) Sampling point location
mm (in.) H20
mm (in.) Hg
Sampling
time,
tain
Total
Clock
time,
2k h
Velocity head
mm (in.) H.,0,
. ( P)
Avg
Flowineter
setting
L/min (ft3/min)
Avg
stack -
°C (°F)
Avg
probe
°C (°F)
Avg
Temperature
sample line
°C (°F)
Avg
readings
flbwmeter box
°C (°F)
;
Avg
container
°C (°F)
Avg
o
o
Quality Assurance Handbook Ml8
-o
-------
FIELD SAMPLING DATA FORM FOR DIRECT INTERFACE SAMPLING
Plant
City
Operator
Date
Barometric press
Initial probe setting
Sampling rate
mm (in.) Hg
°C (°F
Sampling point location
Sample loop volume
Sample loop temp
L/min (cfm) Column temperature:
ml
Run number
Stack dia, mm (in.)
Meter box number
Stack temp
Static press
mm (in.)
Dilution system:
source flow rate
diluent flow rate
diluent flow rate
Dilution ratio
L/min (cfm)
L/min (cfm)
L/min (cfm)
initial /
program rate
final /
Carrier gas flow
/
°C/min
°C/min
°C/min
ml/min
Dilution system check
Final leak check
Vacuum § check
mm (in.) H20
Time of
injection
24 h
Injection
number
Flo*
source
ml/min
nneter(s) s
diluent
ml/min
ettings
diluent
ml/min
stack
°C (°F)
Temperal
probe
°C (°F)
;ure readings
sample line
°C (°F)
injection port
°C (°F)
Quality Assurance Handbook M18-4.2
-------
FIELD SAMPLING DATA FORM FOR ADSORPTION TUBE SAMPLING
Plant
City
Operator
Date
Run number
Stack dia, mm (in.)
Flowmeter calib.(Y) _
Adsorption tube type:
charcoal tube
silica gel
other
Dilution system: (dynamic)
emission flowsetting
diluent flowsetting
Dilution system: (static)
emission flowsetting
Final leak check
Meter box number
Pitot tube (Cp) _
Static press
Adsorption tube number
Average ( P) mm (in.) H20 Vacuum during leak check
m3/min (cfm)
Initial flowineter setting
Average stack temp
mm (in.) H.O
mm (in.) H20 Barometric press
'C (°F) Sampling point location
nm (in.) Hg
Sampling
time,
min
Total
Clock
time,
24 h
Velocity head
mm (in.) H.O,
( P)
Avg
Flowmeter
setting
L/min (ft3/min)
Avg
stack
°C (°F)
Avg
Temperature i
probe, line
°C (°F)
Avg
'eadings
adsorp. tube
°C (°F)
Avg
meter
°C (°F)
Avg
Vacuum
mm (in.) Hg
Avg
o
o
Quality Assurance Handbook MlS-^
-------
ANALYSIS OF METHOD 18 FIELD SAMPLES
Date:
Location:
Analyst:
Plant:
Sample Type:
Type of Calibration Standard:
Number of Standards: Date Prepared:
Target Compound:
Prepared By:
GC Used:
Carrier Gas Used: •
Column Temperatures, Initial:
Sample Loop Volume:
Detector Temp.:
Column Used:
Carrier Gas Flow Rate:
Program Rate:
Loop Temperature:
Auxiliary Gases:
Final:
Inject. Port Temp.:
Calibration Data Standard 1
First analysis/second analysis
Standard concentration (Cact)
Flow rate through loop (ml/min) /
Liquid injection volume (tubes) /
Injection time (24-hr clock) /
Chart speed (cm/min) /
Detector attenuation /
Peak retention time (oin) /
Peak retention time range (min)
Peak area /
Peak area x attenuation factor /
Average peak area value (Y)
Percent deviation from average
Calculated concentration (Cstd)
% deviation from actual (#Dact)
Linear regression equation; slope (m):
Standard 2
Standard 3
y-intercept (b):
Sample Analysis Data
First analysis/second analysis
Sample identification
Interface dilution factor
Flow rate through loop (ml/min)
Liquid injection volume (tubes)
Injection time (24-hr clock)
Chart speed (cm/min)
Detector attenuation
Peak retention time (min)
Peak retention time range (min)
Peak area
Peak area x at ten. factor (AJ/A
Average peak area value (Y)
% deviation from average (%D
Calculated concentration (C )
Sample 1
Sample 2
Sample 3
g
)
' (Y -
/•» I-IT-'P -
^-std Or °s
m
b) '
*VV8 =
-A, - Y
Y
v mot 2n -
x ±uu^ ^ua c t -
td ~ Cact
Y
x 100?
Quality Assurance Handbook M18-5.1
-------
PREPARATION OF STANDARDS BY DILUTION OF GAS CYLINDER STANDARDS
o
Date: Preparer: Purpose:
Cylinder Component: Source:
Component Concentration (X-): ppm Certification Date:
Stage 1 Mixture 1 Mixture 2 Mixture 3
Standard gas flowmeter reading
Diluent gas flowmeter reading
Laboratory temperature ,(°K)
Barometric pressure (Pb) (mm Hg)
Flow rate of cylinder gas (qcl) at
standard conditions (ml/min)
Flow rate of diluent gas (
-------
PREPARATION OF STANDARDS IN TEDLAR BAGS BY GAS AND LIQUID INJECTION
Date:
Preparer:
Purpose:
Organic Compound:
Compound Source:
Gas:
or Liquid:
Compound Purity (P) :
% Compound Mole Weight (M):
Gas Injection Mixture 1 Mixture 2 'Mixture 3
Bag number or identification _ _. _ _ _
Dry gas meter calibration factor (Y) _ . _ _ ••
Final gas meter reading, liters _ _ ,
Initial gas meter reading, liters • _ _
Volume metered (VB), liters _______ _ _
Ambient temperature, °C _ _ _ _
Average gas meter temperature, °C _ _ _
Absolute gas meter temp. (T0), °K _ _ ________
Barometric pressure (Pb), mm Hg ' _ _ _____ _
Average gas meter pressure, mm Hg _ _ _ _
Absolute gas meter press. (PB). mm Hg _ _ _
Gas volume injected (Gv ) , ml _ _ . _
Syringe temperature (T_ ) , °K _ ____ _ _
Absolute syringe pressure (P } , mm Hg _ _ _
Calculated concentration (C.) _ _ _
O t mm~—~^—^ - ««^^— i i !• ......
GV x 103 x
P. x TB
. calc
. corr
100*
Vp x Y
Liquid Injection Mixture 1 Mixture 2 Mixture 3
Bag number or identification _ ___ _ _
Dry gas meter calibration factor (Y) _ _ _
Final gas meter reading- liters _ _ _
Initial gas meter reading, liters _ _ _
Volume metered (VB ) , liters _ _ _ _
Average gas meter temperature, °C _ _ _
Absolute gas meter temp. (TB), °K _ _ _
Barometric pressure (Pb ) , mm Hg •• _ ' _
Average gas meter pressure, mn Hg _ _ _
Absolute gas meter press. (PB)t mm Hg _ _ __ . _
Liquid organic density (p) , ug/ml _ _ _
Liquid volume injected (Lv ) , ttl _ _ __^__ __ _
Calculated concentration (C_ ) • _ _
, , „
C = 6.24 x 104 x
M x V0 x Y x P0
8 corp
100%
Quality Assurance Handbook H18-5-6
-------
DEVELOPMENT OF RELATIVE RESPONSE FACTORS AND RELATIVE RETENTION FACTORS
o
Date:
Target Compound:
Surrogate Compound:
Preparer:
Purpose:
Type of Standard:
Type of Standard:
Target Compound Calibration Data Standard 1
First analysis/verify analysis
Standard concentration
Flow rate through loop (ml/min) /
Liquid injection volume (tubes) /
Injection time (24-hr clock) /
Chart speed (cm/min) /
Detector attenuation /
Peak retention time (tRxl/tRxf) /
Peak retention time range
Peak area /
Peak area x atten. factor (Y1/YX) /
Verification analysis conc.(Cx)
Percent deviation from actual
Caculated retention time (rTxf)
Percent deviation from actual
Linear regression equation; slope (m_):
Standard 2
Standard 3
y- intercept (b) :
O
Surrogate Calibration Data
First analysis/second analysis
Standard concentration
Flow rate through loop (ml/min)
Liquid injection volume (tubes)
Injection time (24-hr clock)
Chart speed (cm/mni)
Detector attenuation
Peak retention time (tRsl/tRg.r)
Peak retention time range
Peak area
Peak area x attenuation factor
Linear regression equation; slope (ms)
Standard 1
Standard 2 Standard
(mf):.
y-intercept (b):
Nonretained peak retention time (tM1/tMf):
Relative Response Factor (F_ ):
Relative Retention Factor (rx/s):
m_
Lx/.
(tRsi - tM1)
x F,
Rx
s td
O
Quality Assurance Handbook M18-5-8
-------
DATA FORM FOR PREPARATION OF LIQUID STANDARDS AND DESORPTION EFFICIENCY SAMPLES
Date:
Preparer:
Purpose:
Organic Compound: _
Compound Source:
Adsorbent Material:
Gas:
or Liquid:
Compound Purity (P):
Batch No:
Jt Compound Mole Weight (M):
Desorption Solvent:
Standards in Solvent
Mixture 1
Mixture 2
ml
Mixture 3
Desorption solvent volume (V,),
Compound spike amount (V0), ul
Organic compound density (p), ug/ul
Standard concentration (Cg), ug/ml
Standards on Adsorbent
Adsorbent amount, g
Compound spike amount (V0), ul
Organic compound density (p), ug/ul
Desorption solvent volume (Ve), ml
Desorption time, min
Standard concentration (CB), ug/ml
Mixture 1 Mixture 2 Mixture 3 Blank
GC Operating Conditions
Injection port temperature, °C
Carrier gas flow rate, ml/min
Column temperature:
Initial, °C
Program rate, °C/min
Final, °C
Chromatographic Results
Injection time, 24-hr clock
Distance to peak, cm
Chart speed, cm/min
Retention time, min
Attenuation factor
Standards in desorption solvent:
Peak area (Ae), area counts
Standards and blank from
adsorbent material:
Peak area (AB and
area counts
Mixture 1 Mixture 2 Mixture 3 Blank
Desorption Efficiency Calculation Mixture 1 Mixture 2 Mixture 3
Desorption Efficiency (DE), %
Vo x p x P
Vs x 1002
DE
A. -
x 100*
Quality Assurance Handbook M18-5-9
-------
CALCULATION FORM FOR GC ANALYSIS BY GAS INJECTION
o
SAMPLE CONCENTRATION
CB = ppm, Pr = . on Hg, T± = . °K,
Pi = . _ mm Hg, Tr = . _ °K, BM8 = 0 . ,
K* = 0 ., F =
C. Pr Tt Fr K
Cc = - = ___ ppm Equation 6-1
PI Tr (1 - BW.K)
'If applicable.
O
Quality Assurance Handbook Ml
o
-------
CALCULATION FORM FOR GC ANALYSIS BY LIQUID INJECTION
SAMPLE VOLUME, DRY BASIS AT STANDARD CONDITIONS
V0
= ._ °K. Bw/ = 0. . K' = 0.
P V
bar o
V.ta.dry s 0.3858 = L ~ Equation 6-2
- B
w.
"If applicable.
DESORPTION EFFICIENCY
Qr= , Qtt= , B =
DE = (Q,. - B)/Qa = o . Equation 6-3
SAMPLE CONCENTRATION
Wp = ug, Wb = ug, Bp = ug.
Bb = ug, V.td = . L, DE = 0. .
K' = 0 .
-------
FIELD AUDIT REPORT
Part A. - To be filled out by organization supplying audit cylinders.
1. Organization supplying audit sample(s) and shipping address
2. Audit supervisor, organization, and phone number
3. Shipping instructions: Name, Address, Attention
o
5.
6.
Guaranteed arrival date for cylinders -
Planned shipping date for cylinders -
Details on audit cylinders from last analysis
d. Audit gas (es) /balance gas..
Low cone.
High cone
Part B. - To be filled out by audit supervisor.
1. Process sampled -
Audit location
2.
3.
Name of individual audit
Audit date
5. Audit Results:
O
d. Measured concentration, ppm
e. Actual audit concentration, ppm
f. Audit accuracy:1
Percent1 accuracy =
Measured Cone. - Actual Cone. x 100
Actual Cone.
Low
cone.
cylinder
High
cone.
cylinder
1 Results of two consecutive injections that meet the sanple analysis
criteria of the test method.
Quality Assurance Handbook Ml8-£
-------
METHOD 18 AUDIT CHECKLIST
Yes
No
Comments
Operation
PRESAMPLING PREPARATION
1. Knowledge of process operations
2. Results of pretest audit (+_ 10% or other value)
3. Calibration of pertinent equipment, in
particular, dry gas meters and other flowmeters
4. Selection and checkout of equipment for proper
sampling and analytical techniques
BAGS - reactivity, condensation, & retention
ADSORPTION TUBES - adsorption & desorption
efficiency
DILUTION SYSTEM - dilution ratio
GC/COLUMN - adequate resolution
GC/DETECTOR - acceptable accuracy & precision
ON-SITE MEASUREMENTS
5- Results of on-site audit (+_ 10% or other value)
6. Sampling system properly assembled
7. Based on pi tot tube check, is proportional
sampling required (more than 10% flow change)
8'. Dilution system check acceptable (if applicable)
9. Sampling system leak check acceptable
10. Proportional sampling properly conducted
11. Constant rate sampling properly conducted
12. Heater systems maintained at proper temperatures
13. Proper number of samples & sampling time
14. GC properly calibrated
15. Duplicate injections had acceptable precision
16. Recording of pertinent process conditions during
sample collection, samples properly identified,
and calculations properly conducted
POSTSAMPLING
17. Results of off-site audit (+, 10% or other value)
18. GC properly calibrated
19. Duplicate injections had acceptable precision <5%
20. Adsorption efficiency acceptable,>9Q% on primary
21. Desorption efficiency accept able,>50% recovery
22. Adequate peak resolution
23. Bags passed reaction check, less than 10% change
24. Bags passed retention check,less than 5# retained
25. Flowmeters recalibration acceptable
26. Temperature sensor recalibration acceptable
COMMENTS
Quality Assurance Handbook M18-8.2
-------
o
o
o
-------
Section No. 3.1?
Date May 31, 1991
Page 1
Section 3.1?
METHOD 25 - DETERMINATION OF TOTAL GASEOUS NONMETHANE ORGANIC EMISSIONS
AS CARBON FROM STATIONARY SOURCES
OUTLINE
Section
SUMMARY
\
METHOD HIGHLIGHTS
METHOD DESCRIPTION
1.
PROCUREMENT OF APPARATUS
AND SUPPLIES
2. CALIBRATION OF APPARATUS
3. PRESAMPLING OPERATIONS
4. ON-SITE MEASUREMENTS
5. POSTSAMPLING OPERATIONS
6. CALCULATIONS
7. MAINTENANCE
8. AUDITING PROCEDURES
9. RECOMMENDED STANDARDS FOR
ESTABLISHING TRACEABILITY
10. REFERENCE METHODS
11. REFERENCES
Documentation
3.17
3-17
3.17.1
3.17.2
3.17-3
3.17.4
3.17.5
3-17.6
3.17.7
3-17.8
3.17.9
3.17.10
3-17.11
Number
of Pages
2
3
25
9
9
14
30
11
2
9
1
25
°1
-------
Section No. 3.1?
Date Hay 31. 1991
Page 2
o
SUMMARY
Method 25 applies to the measurement of volatile organic compounds (VOC) as
total gaseous nonmethane organics (TGNMO) as carbon in source emissions. Organic
particulate matter will interfere with the analysis and, therefore, a particulate
filter is required. The minimum detectable concentration for the method is 50 ppm
as carbon.
When carbon dioxide (C02) and water vapor are present together in the stack,
they can produce a positive bias in the sample. The magnitude of the bias depends
on the concentrations of C02 and water vapor. As a guideline, multiply the C02
concentration times the water vapor concentration, both expressed as volume
percent. If this product does not exceed 100, the bias can be considered
insignificant. For example, the bias is not significant for source emissions
containing 10 percent C02 and 10. percent water vapor, but it would be significant
for a source with VOC emissions ( near the detection limit and with 10 percent C02
and 20 percent water vapor.
An emission sample is withdrawn from the stack at a constant rate through a
heated filter and a chilled condensate trap by means of an evacuated sample tank.
After sampling is completed, the TGNMO are determined by independently analyzing
the condensate trap and sample tank fractions and combining the analytical results.
The organic content of the condensate trap fraction is determined by oxidizing the
NMO to C02 and quantitatively collecting the effluent in an evacuated vessel; then
a portion of the C02 is reduced to CH^, and measured by a flame ionization detector x*~"\
(FID). The organic content of the sample tank fraction is measured by injecting af J
portion of the sample into a gas chromatographic column to separate the NMO fromV—'
carbon monoxide (CO), C02, and CH^; the NMO are oxidized to C02, reduced to CHft,
and measured by an FID. In this manner, the variable response of the FID
associated with different types of organics is eliminated.
This method is not the only method that applies to the measurement of TGNMO.
Costs, logistics, and other practicalities of source testing may make other test
methods more desirable for measuring VOC contents of certain effluent streams.
Proper judgment is required in determining the most applicable VOC test method.
For example, depending upon the molecular weight of the organics in the effluent
stream, a totally automated semicontinuous:; nonmethane organics (NMO) analyzer
interfaced directly to the source may yield accurate results. This approach has
the advantage of providing emission data semicontinuously over an extended time
period.
Direct measurement of an effluent with an FID analyzer may be appropriate with
prior characterization of the gas stream and knowledge that the detector responds
predictably to the organic compounds in the stream. 'If present, methane (CHA)
will, of course, also be measured. The FID can be applied to the determination of
the mass concentration of the total molecular structure of the organic emissions
under any of the following limited conditions: (1) where only one compound is known
to exist; (2) when the organic compounds consist of only hydrogen and carbon; (3)
where the relative percentages of the compounds are known or can be determined, and
the FID responses to the compounds are known; (^) where a consistent mixture of the
compounds exists before and after emission control and only the relative
concentrations are to be assessed; or (5) where the FID can be calibrated against
mass standards of the compounds emitted (solvent emissions, for example).
Another example of direct use of an FID is as a screening method. If there is
enough information available to provide a rough estimate of its accuracy, the FID
analyzer can be used to determine the VOC content of an uncharacterized gas stream.
O
-------
Section No. 3.1?
Date May 31, 1991
Page 3
With a sufficient buffer to account for possible inaccuracies, direct use of an FID
can be a useful tool to obtain the desired measurements without costly exact
determination. In situations where qualitative/quantitative analysis of an
effluent stream is desired or required, a gas chromatographic FID system may apply.
However, for sources emitting numerous organics, the time and expense of this
approach will be formidable.
-------
Section No. 3.17
Date May 31, 1991
Page 4
o
METHOD HIGHLIGHTS
Section 3.17 describes the procedures and specifications for determining
volatile organic compounds as total gaseous nonmethane organics from stationary
sources. An emission sample is withdrawn from the stack at a constant rate
through a heated filter and a chilled condensate trap by means of an evacuated
sample tank. After sampling is completed, the TGNMO are determined by
independently analyzing the condensate trap and sample tank fractions and combining
the analytical results. The organic content of the condensate trap fraction is
determined by oxidizing the NMO to C02 and quantitatively collecting the effluent
in an evacuated vessel; then a portion of the C02 is reduced to CH/, and measured by
an FID. The organic content of the sample tank fraction is measured by injecting a
portion of the sample into a gas chromatographic column to separate the NMO from
carbon monoxide (CO), C02, and CHA; the NMO are oxidized to C02, reduced to CH4,
and measured by an FID. In this manner, the variable response of the FID
associated with different types of organics is eliminated.
On October 3, 1980 (45 FR 65956), EPA published Method 25, "Determination of
Total Gaseous Nonmethane Organic Emissions as Carbon". Shortly after publication,
testers began to report erratic results with the method and suggested a number of
different causes for the imprecision. As a result, EPA began a program to review
the test method in March 1982. The EPA completed the review and proposed revisions
to Method 25, designed to make the njethod simpler, more reliable, and more/*—\
precise. The results of the various studies on Method 25 are presented in thef J
documents listed in Reference Subsection 3-17-H (References 1 through 8). ^-^
On February 12, 1988, several changes were made to Method 25. The studies had
shown that the basic operating principle of Method 25 was sound, but some changes
in equipment design and operating practices would improve the reliability of the
method. These changes can be discussed!by dividing the method into three parts:
Sampling, sample recovery, and analysis.
The major changes in the sampling equipment are the addition of a heated
filter, a redesigned condensate trap, and a different packing material for the
condensate trap. The purpose of the heated filter is to remove organic particulate
matter from the sample and, thus, eliminate a potential source of imprecision. It
is heated to a temperature of 120 °C (2^8 °F). The new trap design is a simple U-
tube which may be more easily and cheaply produced than the previous design. It
also provides a faster and more complete sample recovery than the existing trap
while showing equal collection efficiency. The new packing material is quartz
wool, which, compared to the previously specified stainless steel packing, is more
durable and has improved collection efficiency.
The major changes in the sample recovery are a new oxidation catalyst, a
simplified recovery system, and lower operating temperatures. The new oxidation
catalyst has proven to be very durable and to provide 100 percent oxidation
efficiency for a wide variety of organic compounds at much lower operating
temperatures than the old catalyst. Thejredesigned recovery system has eliminated
some of the tubing and valving and, thus, reduced the potential for sample loss
during recovery and decreased the recovery time. The lower temperatures for
sample recovery will increase the life expectancy of the recovery system materials
and simplify the operation of the system. /*~\
The major change in the sample analysis system is a new separation column for( J
the nonmethane organics analyzer. This new column provides separation of CO, C02,^>—'
and CH^, from a wider range of organic j compounds than the previously specified
column.
-------
Section No. 3.1?
Date May 31, 1991
Page 5
In addition to these major changes, there are a number of minor changes,
particularly in the areas of quality assurance (QA) and calibration.
Collaborative testing of Method 25 has not been performed. However, results
for analysis of performance audit samples have shown that the revised method can
meet the required relative error of +_ 20 percent of the actual concentration of the
audit gas.
The blank data forms may be removed from the Handbook and used in the
pretest, on-site, and posttest operations. The items/parameters that can cause the
most significant errors are designated with an asterisk.
1. Procurement of Apparatus and Supplies
Section 3-17-1 (Procurement of Apparatus and Supplies) gives specifications,
criteria, and design features for the required equipment and materials. This
section can be used as a guide for procurement and initial checks of equipment and
supplies. The activity matrix (Table 1.1) at the end of the section is a summary
of the details given in the text and can be used as a quick reference.
2. Pretest Preparations
Section 3-17-2 (Calibration of Apparatus) describes the required calibration
procedures and considerations for the Method 25 sampling equipment. Required
accuracies for each component are also included. A pretest checklist (Figure 3-1
in Subsection 3-17-3) or a similar form should be used to summarize the calibration
and other pertinent pretest data. The calibration section may be removed along
with the corresponding sections for the other methods and made into a separate
quality assurance reference manual for use by personnel involved in calibration
activities.
Section 3-17-3 (Presampling Operations) provides the tester with a guide for
equipment and supplies preparation for the field test. A pretest preparation form
can be used as an equipment checkout and packing list. Because of the potential
for high blank levels, special attention must be paid in the preparation of the
sampling equipment. Also the tester must ensure that the agency obtains the
required audit samples for the test.
Activity matrices for the calibration of equipment and the presampling
operations (Tables 2.1 and 3-1) summarize the activities.
3- On-Site Measurements
Section3-17-4(On-Site Measurements) contains step-by-step procedures for
sample collection and sample preparation for transport. The on-site checklist
(Figure 4.2, Section 3.17.4) provides the tester with a quick method of checking
the on-site requirements. The revised sampling equipment and procedures were
designed to help eliminate the contamination of the sample for particulate matter
from the source and to provide better collection of condensible organic compounds
in the trap and must be closely followed to provide more precise measurements. The
audit samples are collected during the field sampling phase. Table 4.1 provides an
activity matrix for all on-site activities.
4. Posttest Operations
Section 3-17-5 (Posttest Operations) presents the posttest equipment
procedures and a step-by-step analytical procedure for determination of the total
nonmethane gaseous organics as carbon. Posttest calibration is not required for
any of the sampling equipment. The posttest operations form (Figure 5.1. Section
3.17.5) provides some key parameters to be checked by the tester and laboratory
personnel. The step-by-step analytical procedure description can be removed and
-------
Section No. 3.1?
Date May 31, 1991
Page 6
o
made into a separate quality assurance analytical reference manual for the
laboratory personnel. Initial performance tests of both the condensible organic
recovery system and the NMO analyzer must be performed before the systems are first
placed into operation, after any shutdown of longer than six months, or after any
major modification of the systems. In addition to the initial performance checks,
daily performance checks and calibrations must be performed. Analysis of two audit
samples is required. Strict adherence to Method 25 analytical procedures must be
observed.
Section 3-17-6 (Calculations) provides the tester with the required equations,
nomenclature, and significant digits. Because of the complex nature of the method
and the large number of checks, an example data reporting format is shown. It is
suggested that a calculator or computer be used, if available, to reduce the
chances of calculation error.
Section 3•17-7 (Maintenance) provides the tester with a guide for a
maintenance program. This program is not required, but should reduce equipment
malfunctions. Activity matrices (Tables 5-It 6.1, and 7-1) summarize all
postsampling, calculation, and maintenance activities.
5. Auditing Procedures
Section 3-17-8 (Auditing Procedure) provides a description of necessary
activities for conducting performance and system audits. The performance audit of
the sampling and analytical phase can be conducted using audit gas cylinders*—v
supplied by the Quality Assurance Division, Atmospheric Research and Exposurf J
Assessment Laboratory, U. S. Environmental Protection Agency. The data processing-*—/
procedures and a checklist for a systems audit are also included in this Section.
Table 8.1 is an activity matrix for conducting the performance and system audits.
Section 3-17-9 (Recommended Standards for Establishing Traceability) provides
the primary standard to which the analytical data should be traceable.
6. References
Section 3.17.10 contains the promulgated Method 25 and Section 3.17.11
contains the references cited throughout the text.
O
-------
Section No. 3.17.1
Date May 31, 1991
Page 1
1.0 PROCUREMENT OF APPARATUS AND SUPPLIES
A schematic of the sampling train used for Method 25 is shown in Figure 1.1
and a schematic of the analytical equipment is shown in Figure 1.2.
Specifications, criteria, and/or design features are presented in this section to
aid in the selection of equipment. Many of the sampling train components can be
manufacturer by the tester or other vendors. Procedures and limits (where
applicable) for acceptance checks are also presented. Calibration data generated
in the acceptance checks are to be recorded in the calibration log book.
During the procurement of equipment and supplies, it is suggested that a
procurement log be used to record the descriptive title of the equipment,
identification number (if applicable), and the results of acceptance checks.
The following procedures and descriptions are only provided as guidance to the
tester and may not be requirements of the method for the initial ordering and check
out of the equipment and supplies. The tester should note that many of these
procedures are required at a later step in the sampling and analytical procedures.
It is therefore in the best interest of the sampling and analytical firm that these
procedures or other similar procedures be instituted as routine practice for
checking new equipment and supplies to prevent later problems and/or delays in
test programs. Table 1.1 at the end of this section contains a summary of quality
assurance activities for procurement and acceptance of apparatus and supplies.
1.1 Sampling
The sampling system consists of a heated probe, heated filter, condensate
trap, flow control system, and sample tank (Figure 1.1). The TGNMO sampling
equipment can be constructed from commercially available components and components
fabricated in a machine shop. Complete sampling systems are commercially available
that have been designed to meet all EPA equipment design specifications. The
following equipment is required:
1.1.1 Heated Probe - 6.4-mm (1/4-in.) outside diameter (OD) stainless steel tubing
with a heating system capable of maintaining a gas temperature at the exit end of
at least 129°C (265°F). The probe shall be equipped with a thermocouple at the
exit end to monitor the gas temperature,
A suitable probe is shown in Figure 1.1. The nozzle is an elbow fitting
attached to the front end of the probe while the thermocouple is inserted in the
side arm of a tee fitting attached to the rear of the probe. The probe is wrapped
with a suitable length of high temperature heating tape, and then covered with two
layers of glass cloth insulation and one layer of aluminum foil.
NOTE: If it is not possible to use a heating system for safety reasons, an
unheated system with an in-stack filter is a suitable alternative.
Upon receipt or after construction, visually check the probe for problems,
and plug in the probe heating system to ensure it will heat. Check the
thermocouple at room temperature to ensure it is functional, and check the probe
heating system in conjunction with the entire sampling system as described later in
Subsection 1.1.10. If desired, it may be checked separately by following the
checkout procedures in Subsection 1.1.10 that relate to the sample probe.
1.1.2 Filter Holder - 25-mm (15/l6-in.) inside diameter (ID) Gelman filter holder
or equivalent with stainless steel body and stainless steel support screen with the
Viton 0-ring replaced by a Teflon 0-ring. Upon receipt or after construction,
-------
Section No. 3.17.1
Date May 31. 1991
Page 2
o
REGULATING
VALVE
DUAL RANGE
ROTAMETER
TEMPERATURE
CONTROLLER
THERMOCOUPLE:
I
PURGE VALVE
THERMOCOUPLE
STACK
WALL
MANOMETER
VACUUM PUMP
ROW
CONTROL
VALVE
STAINLESS STEEL
FILTER HOLDER
ROTAMETER P"""!-
HEATED BOX
STAINLESS
STEEL PROBE
SAMPLE
TANK
VALVE
O
CONDENSATE
TRAP
SAMPLE
TANK
Figure 1.1. Method 25 sampling train.
O
-------
Section No. 3-17.1
Date May 31, 1991
Page 3
CALIBRATION STANDARDS
SAMPLE TANK
CARRIER GAS
SAMPLE
INJECTION
LOOP
SEPARATION
COLUMN
CO,CH4.CO2
INTERMEDIATE COLLECTION
VESSEL (CONDITIONED
TRAP SAMPLE)
BACKFLUSH
NONMETHANE
ORGANICS
OXIDATION
CATALYST
REDUCTION
CATALYST
FLAME
IONIZATION
DETECTOR
• HYDROGEN
COMBUSTION
AIR
DATA RECORDER
Figure 1.2. Method 25 analytical equipment.
-------
Section No. 3.17.1
Date May 31, 1991
Page 4
visually check the filter holder for problems, ensure that the unit will properly
hold a filter and tighten, and, then check the filter holder as part of a unit as
described in Subsection 1.1.10. If desired, it may be checked separately by
following the checkout procedures in Subsection 1.1.10 that relate to the filter
holder.
NOTE: Mention of trade names or specific products does not constitute
endorsement by the Environmental Protection Agency.
1.1.3 Filter Heating System - A metal box consisting of an inner and an outer
shell separated by insulating material with a heating element in the inner shell
capable of maintaining a gas temperature at the filter of 121 ^ 3° C (250 +_ 5°F).
A suitable heating box is shown in Figure 1.3. The outer shell is a metal
box that measures 102 mm x 280 mm x 292 mm (4 in. x 11 in. x. 11 1/2 in.), while the
inner shell is a metal box measuring 76 mm x 229 nun x 24l mm (3 in. x 9 in. x 9 1/2
in.). The inner box is supported by 13-mm (1/2-in.) phenolic rods. The void space
between the boxes is filled with fiberfrax insulation which is sealed in place by
means of a silicon rubber bead around the upper sides of the box. A removable lid
made in a similar manner, with a 25-mm (1-in.) gap between the parts, is used to
cover the heating chamber.
The inner box is heated with a 250-watt cartridge heater, shielded by a
stainless steel shroud. The heater is regulated by a thermostatic temperature
controller set to maintain a temperature of 121°C (250°F) as measured by a
thermocouple in the gas line just before the filter. An additional thermocouple is —-
used to monitor the temperature of the gas behind the filter. f j
Upon receipt or after construction, visually check the out-of-stack filter box\_x
for problems, plug in the heater to ensure operation, check the thermocouple at
room temperature to ensure that it is functional, and then check the box as part of
the unit as described in Subsection 1.1.10. If desired, it may be checked
separately by following the checkout procedures in Subsection 1.1.10 that relate to
the filter heating system.
1.1.4 Condensate Trap - 9.5-mm (3/8-in.) OD 316 stainless steel tubing bent into a
U-shape. Exact dimensions are shown in Figure 1.4. The tubing shall be packed
with coarse quartz wool (8 to 15 um), to a density of approximately 0.11 g/cc
before bending. While the condensate trap is packed with dry ice in the Dewar, an
ice bridge nay form between the arms of the condensate trap making it difficult to
remove the condensate trap. This problem can be prevented by attaching a steel
plate between the arms of the condensate trap in the same plane as the arms to
completely fill the intervening space.
Upon receipt or after construction, visually check the condensate trap for
problems, ensure proper fittings, ensure proper packing, and then check the
condensate trap as part of a unit as described in Subsection 1.1.10. If desired,
it may be checked separately by following the checkout procedures in Subsection
1.1.10 that relate to the condensate trap.
1.1.5 Valve - Stainless steel shut-off valve for starting and stopping sample
flow. Upon receipt, visually check the valve for problems and then check the valve
as part of a unit as described in Subsection 1.1.10.
1.1,6 Metering Valve - Stainless steel control valve for regulating the samplef^
flow rate through the sampling train. Upon receipt, visually check the valve f
problems and then check the valve as part of a unit as described in Subsection
1.1.10.
-------
Section No. 3-17.1
Date May 31, 1991
Page 5
VACUUM PUMP
CONNECTOR
SAMPLE
SHUT-OFF
VALVE
25.4
1.0
FIBERFAX
INSULATION
DIMENSIONS: -DEI
in
lijj/ PROBE
CONNECTOR
3.175
0.125
CONDENSATE
TRAP PROBE
BULKHEAD
CONNECTOR
J
PROBE LINE
THERMOCOUPLE
TO TEMPERATURE
CONTROLLER
cn
FILTER HEAT CONDENSATE
TEMPERATURE TRAP PROBE
CONTROLLER CONNECTOR
THERMOCOUPLE THERMOCOUPLE
Figure 1.3. Schematic of out-of-stack filter box.
-------
Section No. 3.17.1
Date May 31, 1991
Page 6
o
DIMENSIONS:
mm
in
0.89
0.375 ^ 0.035
316SS TUBING
316SS NUT
WALL
O
COARSE QUARTZ
WOOL PACKING
2.25
O
Figure 1.4. Condensate trap.
-------
flFTls.
1
Section No. 3.17.1
Date May 31, 1991
Page 7
1.1.7 Rotameter - Glass tube with stainless steel fittings, capable of measuring
sample flow in the range of 60 to 100 cc/min.
Upon receipt, visually check the rotameter for problems and proper range,
ensure proper fittings, and then check it as part of the unit as described in
Subsection 1.1.10. If desired the unit may be checked separately by following the
checkout procedures in Subsection 1.1.10 that relate to the rotameter.
1.1.8 Sample Tank - Stainless steel or aluminum tank with a minimum volume of 4
liters. Upon receipt or after construction, visually check the sample tank for
problems and minimum size, ensure proper fittings, and then check the sample tank
as part of a unit as described in Subsection 1.1.10. If desired, it may be
checked separately by following the checkout procedures in Subsection 1.1.10 that
relate to the sample tank.
1.1.9 Mercury Nanometer or Absolute Pressure Gauge - Capable of measuring
pressure to within 1 mm Hg in the range of 0 to 1,200 mm Hg. Upon receipt or after
construction, visually check the pressure gauge for problems, ensure proper
fittings, proper range, and proper sensitivity, and then check it as part of a
unit as described in Subsection 1.1.10. If desired, it may be checked separately
by following the checkout procedures in Subsection 1.1.10 that relate to the
pressure gauge.
1.1.10 Vacuum Pump - Capable of evacuating a container to an absolute pressure of
10 mm Hg. Upon receipt, visually check the vacuum pump for problems. Ensure
proper fittings, attach the pump to a vacuum gauge, and determine if the pump is
capable of evacuating to an absolute pressure of 10 mm Hg. Check it as part of a
unit as described below. If desired, it may be checked separately by following the
checkout procedures below that relate to the pump.
1. Calibrate all thermocouples as described in Section 3-17-2. If the
thermocouples are not within 3°C (5°F) of the true temperature, reject or
repair and recalibrate.
2. Calibrate the rotameters as described in Section 3.17.2. If the
rotameters cannot determine the flowrate to within 10 percent of the
actual flowrate over the indicated range, reject or repair and then
recalibrate.
3. Calibrate and leak check the sample tank as described in Section 3.17.2.
The tank is acceptable if no change in tank vacuum is noted over a 1 hour
period.
4. With the sample tank evacuated, assemble the sampling train (including
placing a filter in the filter holder) as shown in Figure 1.1, with the
exception that is not necessary to use dry ice to cool the condensate
trap. Plug the probe tip and verify that the tank valve is closed. Turn
on the vacuum pump, and evacuate the sampling system from the probe tip to
the sample tank valve to an absolute pressure of 10 mm Hg or less. Close
the purge valve, turn off the pump, wait a period of 5 minutes, and re-
check the indicated vacuum. The method allows a leak rate of 1 percent
of the sampling rate, but for this initial check of new equipment it is
recommended that the criteria be increased to no noticeable leak. If a
leak is noted, find the source of it and reject, repair, or replace the
component(s) and repeat the leak check until satisfactory results are
obtained.
5. Release the vacuum and then unplug the probe tip. Set the probe
temperature controller to 129°C (265°F) and the filter temperature
-------
Section No. 3.17-1
Date May 31, 1991
Page 8
controller to 121°C (250°F). Allow the probe and filter to heat for
about 30 minutes. Close the sample valve, open the purge valve, start the
vacuum pump, and set the flow rate between 60 and 100 cc/min. If the
purge system draws gas through the probe and filter at the desired rate,
the purge system is acceptable. If the purge system does not provide the
desired flow rate, repair or replace the system or problem component and
repeat the check.
6. When the temperature at the exit ends of the probe and filter are within
the specified range, close the purge valve and stop the pump. Open the
sample valve and the sample tank valve. Using the flow control valve, set
the flow through the sample train to the maximum rate that would be
normally used (i.e., 100 cc/min). Operate the system for about 30
minutes, adjusting the flow rate as necessary to maintain a constant rate
(+10 percent). The temperature of the probe and filter must remain in
the specified range and the flowrate should be adjustable. If the flow
rate and temperatures can be maintained in the proper range, conduct
another leak check as described above in Step 4. If the flow rate and/or
temperatures cannot be maintained in the proper range, repair or replace
system or problem component(s) and repeat sampling check and leak checks.
1.2 Analysis
The analysis equipment consists primarily of an organic condensate recovery
system for processing sample traps and a nonmethane organic (NMO) analyzer for
analysis of C02 and NMO. The recovery system can be constructed from commercially
available components and components fabricated in a machine shop. The NMO analyzer
can be purchased as a commercial unit or assembled in the laboratory by
modification of a packed column gas chromatograph.
1.2.1 Condensate Recovery Apparatus - The system for the recovery of the organics
captured in the condensate trap consists of a heat source, oxidation catalyst,
nondispersive infrared (NDIR) analyzer, and an intermediate collection vessel
(ICV). Figure 1.5 is a schematic of a typical system. The system shall be capable
of proper oxidation and recovery, as specified in Section 3.17.5- The following
major components are required:
Heat Source - Sufficient to heat the condensate trap (including connecting
tubing) to a temperature of 200°C. A system using both a heat gun and an electric
tube furnace is recommended. Upon receipt, visually check the heat gun and/or
electric tube furnace for any defects. Check any device for heating before use to
determine if a condensate trap temperature of 200°C can be achieved.
Heat Tape - Sufficient to heat the connecting tubing between the water trap
and the oxidation catalyst to 100°C. Upon receipt, visually check the heating tape
for any defects and test to ensure heating of the connecting tubing to 100°C.
Oxidation Catalyst - A suitable length of 9.5-mm (3/8-in.) OD Inconel 600
tubing packed with 15 cm (6 in.) of 3-2-mm (1/8-in.) diameter 19 percent chromia on
alumina pellets. The catalyst material is packed in the center of the catalyst
tube with quartz wool packed on either end to hold it in place. The catalyst tube x—-v
shall be mounted vertically in a 650°C tube furnace. After construction, visually ( j
check the catalyst tube for problems and test the tube heater to ensure heating to V '
650°C.
o
-------
Section No. 3.17.1
Date May 31, 1991
Page 9
FLOW METERS
\
HEAT TRACE (100°C)
SAMPLE
RECOVERY
VALVE
FLOW
CONTROL
VALVE
SYRINGE PORT
VACUUM PUMP
Figure 1.5- Condensate recovery system.
-------
Section No. 3.17.1
Date May 31, 1991
Page 10
o
Water Trap - Leak proof, capable of removing moisture from the gas stream.
Upon receipt or after construction, visually check the water trap for defects,
especially leakage problems.
Syringe Port - A 6.4-mm (1/4-in.) OD stainless steel tee fitting with a rubber
septum placed in the side arm. After construction, visually check the syringe port
for defects, especially leakage around the rubber septum.
NDIR Detector - Capable of indicating C02 concentration in the range of 0 to 5
percent, to monitor the progress of combustion of the organic compounds from the
condensate trap. Upon receipt, visually check the NDIR detector for defects. Zero
and span the analyzer according to the manufacturer's directions.
Flow-Control Valve - Stainless steel, to maintain the trap conditioning system
near atmospheric pressure. Upon receipt, visually check the flow control valve for
defects.
Intermediate Collection Vessel (ICV) - Stainless steel or aluminum, equipped
with a female quick connect. Tanks with nominal volumes of at least 6 liters are
recommended. Upon receipt, visually check the ICV for defects.
Determine the ICV volume by weighing it while empty and then filling it with
deionized distilled water; weigh to the nearest 5 g and record the difference
between the full and empty weights as the tank volume in ml. Alternatively.
measure the volume of water used to fill the tank to the nearest 5 ml«
An alternative to using the rigid containers is the use of flexible bags
of Tedlar or Teflon film. However, the calculations given in Section 3-17-6 are no
longer appropriate since the gas volumes must be measured directly. It is the
responsibility of the tester to apply calculations which are consistant with
directly measured gas volumes.
Mercury Nanometer or Absolute Pressure Gauge - Capable of measuring pressure to
within 1 mm Hg in the range of 0 to 1,200 mm Hg. Upon receipt, visually check the
manometer or pressure gauge for defects and proper operating range and precision.
Syringe - 10-ml gas-tight, glass syringe equipped with an appropriate needle.
Upon receipt, visually check the syringe for defects and proper volume.
1.2.2 NMO Analyser - The NMO analyzer is a gas chromatograph (GC) with backflush
capability for NMO and C02 analysis and is equipped with an oxidation catalyst,
reduction catalyst, and FID. Figures 1.6 and 1.7 are schematics of a typical NMO
analyzer. This semicontinuous GC/FID analyzer shall be capable of: (1) separating
CO, C02, and CH4 from NMO; (2) reducing the C02 to CH4, and quantifying as CH4; and
(3) oxidizing the NMO to C02, reducing the C02 to CHa and quantifying as CH^,
according to Section 3-17-5- The NMO analyzer consists of the following major
components:
Oxidation Catalyst - A suitable length of 9.5-mm (3/8-in.) OD Inconel 600
tubing packed with 5-1 cm (2 in.) of 19 percent chromia on 3-2-mm (1/8-in.) alumina
pellets. The catalyst material is packed in the center of the tube and supported
on either side by quartz wool. The catalyst tube must be mounted vertically in
650° C furnace. After construction, visually inspect the oxidation catalyst f
defects and ensure that the tube furnace is capable of heating to 650°C.
-------
Section No. 3.17.1
Date May 31, 1991
Page 11
CARRIER GAS
CALIBRATION STANDARDS
SAMPLE TANK—>
i
SAMPLE
INJECTION
LOOP
INTERMEDIATE COLLECTION
<— VESSEL (CONDITIONED
TRAP SAMPLE)
SEPARATION
COLUMN
BACKFLUSH
CO, CH4, CO2
NONMETHANE
ORGANICS
OXIDATION
CATALYST
REDUCTION
CATALYST
<—HYDROGEN
1
FLAME
lONEATION
DETECTOR
COMBUSTION
AR
DATA RECORDER
Figure 1.6. Simplified schematic of nonmethane organic (NMO) analyzer.
-------
Section No. 3-17.1
Date May 31, 1991
Page 12
o
COLUMN OVEN
REDUCTION
CATALYST
OXIDATION
CATALYST
H2 AIR
Figure 1.7. Nonmethane organic (NMO) analyzer.
O
-------
Section No. 3.17-1
.Pate May 31. 1991
Page 13
Reduction Catalyst - A 7.6-cm (3-in.) length of 6.4-mm (1/4-in.) OD Inconel
tubing fully packed with 100-mesh pure nickel powder. The catalyst tube must be
mounted vertically in a 400°C furnace. After construction, visually inspect the
reduction catalyst for defects and ensure that the tube furnace is capable of
heating to 400°C.
Separation Column(s) - A 30-cm (1-ft) length of 3.2-mm (1/8-in.) OD stainless
steel tubing packed with 60/80 mesh Unibeads is followed by a 6l-cm (2-ft) length
of 3.2-mm (1/8-in.) OD stainless steel tubing packed with 60/80 mesh Carbosieve G.
The Carbosieve and Unibeads columns must be baked separately at 200°C with carrier
gas flowing through them for 24 hours before initial use. The columns should then
be connected to each other with a 1/8-inch stainless steel union. The column
series should be connected to the sample injection valve so that the sample loop
contents will be injected onto the head of the Unibeads IS column.
Sample Injection System - A 10-port GC sample injection valve fitted with a
sample loop properly sized to interface with the NMO analyzer (1-cc loop
recommended). Upon receipt, visually inspect the sample injection system for
defects and check for proper number of ports and valve fitting size for the
connecting tubing used (1/16- or 1/8-in.).
FID - An FID meeting the following specifications is required:
1. Linearity - A linear response (+_ 5 percent) over the operating range as
demonstrated by the procedures established in Section 3«17'5-
2. Range - A full scale range of 10 to 50,000 ppm CH4. Signal attenuators
shall be available to produce a minimum signal response of 10 percent of
full scale.
Data Recording System - Analog strip chart recorder or digital integration
system compatible with the FID for permanently recording the analytical results.
Upon receipt, visually inspect the data recording system for defects and test
according to manufacturer's instructions.
1.2.3 Other Analysis Apparatus -
Barometer - Mercury, aneroid, or other barometer capable of measuring
atmospheric pressure to within 1 mm Hg. Upon receipt, visually check the barometer
for defects.
Thermometer - Capable of measuring the laboratory temperature to within 1°C.
Upon receipt, visually check the thermometer for defects.
Vacuum Pump - Capable of evacuating to an absolute pressure of 10 mm Hg or
less. Upon receipt, visually check the vacuum pump for defects and test to ensure
capability to reach proper vacuum
Syringes - 10 ul and 50 ul liquid injection syringes. Upon receipt, visually
check syringes for defects and proper volume.
Liquid Sample Injection Unit - 316 stainless steel U-tube constructed as shown
in Figure 1.8 for performing condensible organic recovery efficiency tests. After
construction, visually check the unit for problems, especially leakage around the
rubber septum.
-------
Section No. 3.17.1
Date May 31, 1991
Page 14
CONNECTING T
INJECTION
SEPTUM
o
CONNECTING ELBOW
FROM
CARRIER
DIMENSIONS:
TO
CATALYST
O
316SS TUBING
Figure 1.8. Liquid sample injection unit.
O
-------
Section No. 3.17.1
Date May 31, 1991
Page 15
1.3 Reagents and Other Supplies
Unless otherwise indicated, all reagents should meet the specifications of the
Committee on Analytical Reagents of the American Chemical Society (ACS); otherwise,
use the best available grade.
1.3.1 Sampling - The following are required for sampling:
Crushed Dry Ice - Crushed dry ice is needed to cool the condensate (U-tube)
trap during sampling for better collection of organics and to keep it cold until
analysis. There are no specifications on the dry ice.
Coarse Quartz Wool - Coarse quartz wool, 8 to 15 urn in size, is needed to pack
the condensate (U-tube) traps in the laboratory for sampling. The packing should
not be conducted in the field prior to testing since a trap packed with new quartz
wool must be taken to 300° C and then blank checked prior to its use in a field
test. Upon receipt, check the specifications of the quartz wool. If the proper
wool has been sent, it should be acceptable. If the specifications are not met,
reorder the proper item.
Filters - Glass fiber filters, without organic binder are needed to remove
organic particulate matter from the gas stream during sample collection.
Typically, filters used for Method 5 tests will be satisfactory, if no organic
binders are present. If organic binders are present, they may be released during
testing and positively bias the results. If the tester is not certain about the
presence of organic binders in a glass fiber filter, it should be placed in a
furnace at 300°C for 2 hours which will remove any organic binders present. This
procedure, however, may make the filter more brittle resulting in a greater need
for caution in handling. A check on the amount of organic binder lost can be
determined by weighing the filter both before and after heating. If a significant
weight loss 'occurs (1 mg per filter), the filters probably contain organic binders.
They may still be used, but it is recommended that another type of filter be
ordered and checked in the same manner, since removing the binders with heat may
make the filters too brittle to use safely.
1.3.2 NMO Analysis - Several gases are needed for NMO analysis depending on the
exact analyzer used for analysis. It is critical that all gases meet the require-
ments for background contamination, to ensure that a low background level is
present during sample analysis. The following gases are needed for NMO analysis:
Carrier Gases - Depending on the exact NMO analyzer, two carrier gases will be
needed for analysis. Typically zero grade helium (He) and zero grade oxygen (0_)
containing less than 1 ppm C02 and less than 0.1 ppm C as hydrocarbon will be
required. Upon receipt, check the label for manufacturer's specifications. If the
gases do not meet the above specifications, they should be returned to the
supplier, and new gases obtained and checked.
Fuel Gas - Typically zero grade hydrogen (H2) cylinder gas is needed as a fuel
gas. The hydrogen should be 99-999 percent pure. Upon receipt, check the label
and manufacturer's specifications. If the gas does not meet these specifications,
Itllllreturn it to the supplier, and obtain and check a new cylinder.
-------
Section No. 3-17.1
Date May 31, 1991 x~
Page 16 r A
Combustion Gas - Zero grade air or 02 (as required by the GC detector) is
needed. Upon receipt, check the specifications. If the gas does meet the required
specifications, return it to the supplier and obtain new gas and recheck.
1.3.3 Condensate Recovery - Two gases are needed for condensate recovery. It is
critical that all gases meet the requirements for background contamination, to
ensure that a low background level is present during sample analysis. The
following gases are needed for condensate recovery:
Carrier Gas - Zero grade air, containing less than 1 ppm C as hydrocarbons is
needed as a carrier for purging the C02 from the trap into the sample tank and then
purging the sample from the trap during the oxidation step into an intermediate
collection vessel. Upon receipt, the manufacturer's specifications should be
checked and the gases analyzed for background levels as described in Section
3.17.5. If the gas does not meet the requirements, it should be replaced and the
new gas checked.
Auxiliary Oxtdant - Zero grade 02 , containing less than 1 ppm C as
hydrocarbons may be needed during the oxidation of the condensate trap sample.
Upon receipt, the gas should be checked as described above. Zero grade air may be
used instead of zero grade 02 if the condensible organic recovery efficiency test
(described in Section 3-17-5) can be passed.
1.3^ Condensate Recovery Performance - The following liquid reagents are needed:
Hexane - ACS grade hexane is needed for liquid injection into the liquid
sample injection unit of the condensate recovery system to conduct the condensible
organic recovery efficiency test. Upon receipt, check the container of hexane to
ensure that the ACS grade specifications are met. If they are not met, return it
to the supplier, obtain a new container, and recheck.
••1, ». • , ;
Decane - ACS grade decane is needed for liquid injection into the liquid
sample injection unit of the condensate recovery system to conduct the condensible
organic recovery efficiency test. Upon receipt, check the container of decane to
ensure that the ACS grade specifications are met. If they are not met, return it
to the supplier, obtain a new container, and recheck.
1.3•5 Calibration Gases for Analysts - The concentrations of all calibration gases
should be traceable to National Institute for Standards and Technology (NIST)
Standards. For those calibration gases that have corresponding gaseous NIST
standards (i.e., propane and carbon dioxide), traceability should be established
via the EPA's Revised Traceability Protocol No. 1 (Reference 9). For the remaining
calibration gases, traceability should be established to gravimetric NIST
standards. Traceability to NIST is necessary because some calibration gases with
certificates of analysis have shown significant errors when they were compared with
NIST standards. Specialty gas manufacturers should certify the accuracy of their
calibration gases.
Revised Traceability Protocol No. 1 compares the concentrations of calibration
gases to those of gaseous NIST Standard Reference Materials (SRMs) or to those of
gaseous NIST/EPA Certified Reference Materials (CRMs), which are accepted as X~N
equivalent to SRMs (Reference 10). Although explicit accuracy specifications for C j
these EPA protocol gases do not exist, accuracy assessments by EPA have found that ^-—'
many EPA protocol gases are accurate to within 2 percent and that most are accurate
-------
Section No. 3.17-1
Date May 31, 1991
Page 1?
to within 5 percent (Reference 11). EPA protocol gases may be purchased from most
specialty gas manufacturers.
For all calibration gases, the manufacturer must recommend a maximum shelf
life (i.e., the length of time during which the gas concentration is not expected
to change by more than 5 percent from its certified value). EPA protocol gases
have a certification period of 18 months, after which they should be recertified.
Specialty gas manufacturers should be able to produce stability data to support the
maximum shelf life recommendation. The data should be for the same compound,
balance gas, and approximate concentration as requested.
Do not store the calibration gas cylinders in areas subject to extreme
temperature changes. Before each calibration, check the pressure of the
calibration gas in the cylinder and replace any cylinders with a pressure less than
1500 kilopascals (or 200 Ib/sq. in.).
The following calibration gases are required.
Oxidation Catalyst Efficiency Check Calibration Gas - A calibration gas with a
nominal concentration of 1 percent methane in air is required for the oxidation
catalyst efficiency check. Upon receipt of the calibration gas, check its
certificate of analysis to ensure that the correct concentration has been sent.
Verify its certified concentration against calibration gases currently being used
for analysis. Be certain that balance gas differences do not cause measurement
errors in the analyzer used for verification. The verified concentration should
agree within 2 percent of the certified concentration. If 2 percent agreement is
obtained, the certified concentration can be used. If the agreement is between 2
and 5 percent, the verified concentration can be used. If the agreement is greater
than 5 percent, first inspect the analyzer for malfunction and reverify the
calibration gas. If the reverified agreement is also greater than 5 percent,
consult with the specialty gas manufacturer about replacement of the calibration
gas.
The organic condensate recovery system oxidation catalyst efficiency check
compares the concentrations of the methane and C02 calibration gases. The
concentrations of the two calibration gases must agree within 2 percent. Be
certain to compare the calibration gases on a ppm carbon basis. If such agreement
is not obtained, the problem may lie in the catalyst or in the calibration gases.
Inspect or replace the catalyst and reverify the calibration gases' concentrations
before repeating the check. Consult with the specialty gas manufacturer about
replacement of the calibration gases if the lack of agreement persists.
7/WO Response Linearity and Calibration Gases - Three calibration gases with
nominal concentrations of 20, 200, and. 3»000 ppm propane in air are required for
the NMO linearity check and to determine the calibration response factor. Upon
receipt of the calibration gases, check their certificates of analysis to ensure
that the correct concentrations have been sent. Verify the certified
concentrations against SRMs using revised Traceability Protocol No. 1 or against
calibration gases currently being used for analysis as described in Section 3-17-5-
Be certain that balance gas differences do not cause measurement errors in the
analyzer used for certification. The verified concentration should agree within 2
percent of the certified concentration. If 2 percent agreement is obtained, the
certified concentration can be used. If the agreement is greater than 5 percent,
first inspect the analyzer for malfunction and reverify the calibration gas. If
the reverified agreement is also greater than 5 percent, consult with the specialty
gas manufacturer about replacement of the calibration gas.
-------
Section No. 3.1?
Date May 31, 1991
Page 18
•'O
C02 Response Linearity and Calibration Gases - Three calibration gases with
nominal concentrations of 50 ppm, 500 ppm, and 1 percent carbon dioxide in air are
required to determine the overall mean C02 response factor. Upon receipt of the
calibration gases, check their certificates of analysis to ensure that the correct
concentrations have been sent. Verify the certified concentrations against SRMs
using revised Traceability Protocol No. 1 or against calibration gases currently
being used for analysis as described in Section 3-1?-5- Be certain that balance
gas differences do not cause measurement errors in the analyzer used for
certification. For example, carbon dioxide in air response factors will differ
from carbon dioxide in nitrogen response factors for nondispersive infrared
analyzers due to pressure broadening effects. The verified concentration should
agree within 2 percent of the certified concentration. If 2 percent agreement is
obtained, the certified concentration can be used. If the agreement is between 2
and 5 percent, the verified concentration can be used. If the agreement is greater
than 5 percent, first inspect the analyzer for malfunction and reverify the
calibration gas. If the reverified agreement is also greater than 5 percent,
consult with the specialty gas manufacturer about replacement of the calibration
gas.
The analyzer linearity check and NMO calibration requires 10 percent agreement
between the propane and C02 calibration gases. Be sure to compare the calibration
gases on a ppm carbon basis. If such agreement cannot be obtained, inspect the
analyzer for malfunction and reverify the calibration gases. Consult with the^^
specialty gas manufacturer about replacement of the calibration gases if the lacfA
of agreement persists. V_>X
NMO Analyser System Performance Check Calibration Gases - The four following
calibration gases are needed for the NMO analyzer system check:
1. Propane Mixture - A calibration gas with nominal concentrations of 50 ppm
carbon monoxide, 50 ppm methane, 2 percent carbon dioxide, and 20 ppm
propane in air is required for the NMO analyzer system performance check.
Upon receipt of the calibration gas, check its certificate of analysis to
ensure that the correct concentrations have been sent. Replace the
calibration gas if the purchase specifications have not been met. If the
specifications have been met, conduct the NMO calibration check. The
response factor for this calibration gas should be within 5 percent of the
overall mean response factor for the propane in air calibration gases. If
such agreement cannot be obtained, inspect the equipment for malfunction
and verify the concentrations of the components in the calibration gas.
Consult with the specialty gas manufacturer about replacement of the
calibration gas if the lack of agreement persists.
2. Hexane Calibration Gas - A. calibration gas with a nominal concentration of
50 ppm hexane in air is required for the NMO analyzer system performance
check. Upon receipt of the calibration gas, inspect its certificate of
analysis to ensure that the correct concentration has been sent. Replace
the calibration gas if the purchase specifications have not been met. If
the specifications have been met, conduct the NMO calibration check. The
response factor for this calibration gas should be within 5 percent of the
overall mean response factor for the propane in air calibration gases. Be,...
sure to compare the calibration gases on a ppm carbon basis. If suq A
agreement cannot be obtained, inspect the equipment for malfunction and, i\ J
possible, verify the concentration of the calibration gas with an analyzer
employing a different analytical principle (e.g., gas chromatography).
Consult with the specialty gas manufacturer about replacement of the
-------
Section No. 3.17.1
Date May 31, 1991
calibration gas if the lack of agreement persists.
Toluene Calibration Gas - A calibration gas with a nominal concentration of
20 ppm toluene in air is required for the NMO analyzer system performance
check. Upon receipt of the calibration gas, check its certificate of
analysis to ensure that the correct concentration has been sent. Replace
the calibration gas if the purchase specifications have not been met. If
the specifications have been met, conduct the NMO calibration check. The
response factor for this calibration gas should be within 5 percent of the
overall mean response factor for the propane in air calibration gases. Be
sure to compare the calibration gases on a ppm carbon basis. If such
agreement cannot be obtained, inspect the equipment for malfunction and, if
possible, verify the concentration of the calibration gas with an analyzer
employing a different analytical principle (e.g., gas chromatography) .
Consult with the specialty gas manufacturer about replacement of the
calibration gas if the lack of agreement persists.
Methanol Calibration Gas - A calibration gas with a nominal concentration
of 100 ppm methanol in air is required for the NMO analyzer system
performance check. Upon receipt of the calibration gas, check its
certificate of analysis to ensure that the correct concentration has been
sent. Replace the calibration gas if the purchase specifications have not
been met. If the specifications have been met, conduct the NMO calibration
check. The response factor for this calibration gas should be within 5
percent of the overall mean response factor for the propane in air
calibration gases. Be sure to compare the calibration gases on a ppm
carbon basis. If such agreement cannot be obtained , inspect the equipment
for malfunction and, if possible, verify the concentration of the
calibration with an analyzer employing a different analytical principle
(e;g., gas chromatography). Consult the specialty gas manufacturer about
replacement of the calibration gas if the lack of agreement persists.
Note; Little is currently known about the stability of calibration gases
containing methanol in air.. Special attention should be given to the
stability of this calibration gas.
-------
Section No. 3.17.1
Date May 31, 1991
Page 20
Table 1.1. ACTIVITY MATRIX FOR PROCUREMENT OF APPARATUS AND SUPPLIES
o
Apparatus
Sampling
Heated probe
Filter holder
Filter heating
system
Condensate trap
Valve
Metering valve
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Stainless steel cap-
able of heating to
129°C (365°F) at
exit end
Upon receipt, check
heater and thermo-
meter with system
Repair or
replace, and
recheck
Stainless steel with
inside diameter of
25 mm (15/16 in.)
and Teflon 0-ring
Check unit to ensure
that filter is held
properly
Repair or
replace, and
recheck
Metal box consisting
of an inner and an
outer shell separ-
ated by fiber fran
insulation with a
heater capable of
maintaining a gas
temperature of 121°C
+3°C (250 +5°F)
Visually check, ensure
heater is operational,
check thermocouple at
room temperature,
check box as part of
the unit (Subsection
1.1.10)
Repair or
replace and
recheck
O
9.5 mm (3/8 in.)
OD 316 stainless
steel tubing U-tube
shaped, packed with
coarse quartz wool
Visually check, ensure
proper fittings,
proper packing, check
as part of unit
Repair or
replace and
recheck
Stainless steel
shut-off valve
Visually check valve,
check as part of unit
Repair or
replace and
recheck
Stainless steel
control valve
Visually check valve,
check as part of unit
Repair or
replace and
recheck
Rotameter
(Continued)
Glass tube with
stainless steel
fittings, capable
of measuring sample
flow of 60 to 100
cc/min
Visually check, ensure
proper range and
proper fittings, check
as part of unit
Reject or
repair, then
recalibrate
O
-------
Section No. 3.1?.l
Date May 31. 1991
Page 21
Table 1.1 (Continued)
Apparatus
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Sample tank
Stainless steel or
aluminum tank with
a minimum volume of
4 liters
Visually check, ensure
proper fittings, min-
imum size, check as
part of unit
Repair or
replace and
recheck
Mercury manometer
or absolute
pressure gauge
Capable of measur-
ing pressure to
within 1 mm Hg. in
;the range of 0 to
1,200 mm Hg
Visually check, ensure
proper fittings,
range, sensitivity,
check as part of unit
Determine
correction
factor or
reject
Vacuum pump
Capable of evacu-
ating to an absolute
pressure of 10 mm
Hg
Visually check, ensure
proper fittings, de-
termine for evacuating
10 mm Hg, check as
part of unit
Repair or
replace and
recheck
Analysis
Condensate
recovery apparatus
1. Heat source
2. Heat tape
3. Oxidation
catalyst
4. Water trap
5- Syringe port
(Continued)
Sufficient to heat
condensate trap to
200° C
Sufficient to heat
connecting tubing
to 100°C
9.5 mm (3/8 in.) OD
Inconel 600 tubing
packed with 15 cm
(6 in.) of 3-2 cm
(1/8 in.) diameter
19 percent chromia
or alumina pellets
Leak proof, capable
of removing moisture
from gas stream
6.4 mm (1/4 in.) OD
stainless steel
Visually check, con-
duct heat check
Visually check, con-
duct heat check
Visually check
Repair or
replace and
recheck
Replace and
recheck
Repair or
replace and
recheck
Visually check, con-
duct leak check
Visually check
Repair or
replace and
recheck
Repair or
replace and
recheck
-------
Table 1.1 (Continued)
Section No. 3-17.1
Date May 31, 1991
Page 22
o
Apparatus
6. NDIR detector
7. Flow-control
valve
8. Intermediate
collection
vessel
9. Mercury mano-
meter or
absolute pres-
sure gauge
10. Syringe
NMO Analyzer
1. Oxidation
catalyst
2. Reduction
catalyst
Acceptance limits
Capable of indicat-
ing C02 concentra-
tion in the range of
0 to 5 percent
Stainless steel
Stainless steel or
aluminum, nominal
volume of 6 liters
Capable of measur-
ing pressure to
within 1 mm Hg in
the range of 0 to
1,200 mm Hg
10 ml gas-tight,
glass
Suitable length of
9-5 ran OD Inconel
600 tubing packed
with 5.1 cm of 19
percent chromia on
3.2 mm alumina
pellets
7.6 cm length of 6.4
mm OD Inconel tubing
packed with 100 mesh
pure nickel powder
Frequency and method
of measurement
Visually check, check
pressure rating
Visually check, ensure
proper volume
Visually check, ensure
ensure proper sensi-
tivity
Visually check
Visually check
Visually check
Action if
requirements
are not met
Repair or
replace and
recheck
Repair or
replace and
recheck
Repair or
replace and
recheck
Determine
correction
factor or
reject
Repair or
replace and
recheck
O
Replace and
recheck
Replace and
recheck
(Continued)
O
-------
Section No. 3.17-1
Date May 31, 1991
Page 23
Table 1.1 (Continued)
Apparatus
Separation
column
Sample injec-
tion system
5. FID
6. Data recording
system
Acceptance limits
30 cm length of 3-2T
mm OD stainless
steel tubing packed
with 60/80 mesh
Unibeads IS followed
by 6l cm length of
3.2 mm OD stainless
steel tubing packed
with 60/80 mesh
Carbosieve G
10-port GC sample
injection valve
fitted with a
sample loop
A linear response
of +_ 5 percent over
operating range of
10 to 50,000 ppm
CHA, minimum signal
response of 10 per-
cent of full scale
Analog strip chart
compatible with FID
Frequency and method
of measurement
Visually check
Visually check, sample
loop should be of
proper size to inter-
face with NMO analyzer
Upon receipt use
procedure established
in Section 3-17.5
Upon receipt check as
recommended by manu-
facturer
Action if
requirements
are not met
Replace and
recheck
Repair or
replace and
recheck
Return to
manufacturer
or repair
and recheck
Repair or
return to
manufacturer
Reagents and
other supplies
Sampling
1. Crushed dry
ice
2. Coarse quartz
wool
3. Filters
(Continued)
No specifications
8 to 15 urn in size
Glass fiber filters
without organic
binders
None
Upon receipt check
manufacturer's speci-
fications
Check for presence of
organic binders
None
Return to
manufacturer
Return to
manufacturer
' J "7
,' -v '\
/ < A) .
-------
Table 1.1 (Continued)
Section No. 3-17-1
Date May 31, 1991
Page 24
o
Apparatus
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
NMO analysis
1. Carrier gas
2. Fuel gas
3- Combustion gas
As specified by
owner's manual, less
than 1 ppm C02 and
0.1 ppm C as hydro-
carbon
As specified by
owner's manual
As specified by
owner's manual
Upon receipt check
label and manufactur-
er's specifications
Upon receipt check
label and manufac-
turer's specifications
Upon receipt check
label and manufactur-
er's specifications
Return to .
supplier and
check new gas
Return to
supplier and
check new gas
Return to
supplier and
check new gas
O
Condensate
analysis
1. Carrier gas
2. Auxiliary 0.,
3. Hexane
. Decane
Zero grade air con-
taining less than
1 ppm C as hydro-
carbon
Zero grade 02 con-
taining less than
1 ppm C as hydro-
carbon
ACS grade hexane
ACS grade decane
Same as above
Same as above
Visually check to
ensure ACS grade
Same as above
Same as above
Same as above
Return to manu-
facturer and
check new reagent
Same as above
(Continued)
O
-------
Table 1.1 (Continued)
Section No. 3-17.1
Date May 31, 1991
Page 25
Apparatus
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Calibration gases
for analysis
1. Oxidation
catalyst
efficiency
check calibra-
tion gas
2. FID linearity
and NMO cali-
bration gases
3. C02 calibra-
tion gases
Gas mixture with
nominal concentra-
tion of 1 percent
methane
Three gas mixture
standards with
nominal propane
concentrations of
20, 200 and 3000 ppm
Three gas mixture
standards with
nominal C02 concen-
trations of 50, 500,
and 1 percent in air
Visually check
specifications
Check specifications,
compare against
NBS-SRM propane or
previous calibration
gas
Check specifications,
compare against
previous calibration
gases or NBS-SRM
No. 2622 (2% C02 in
N2) diluted with 20%
zero oxygen
Return to
supplier, order
from new
supplier
Return to
manufacturer
Return to
manufacturer
NMO analyzer
system check
calibration gas
1. Propane
mixture
2. Hexane gas
standard
3- Toluene gas
standard
4. Methanol gas
standard
Gas mixture contain-
ing (nominal) 50 ppm
CO, 50 ppm CHA, 2%
C02, and 20 ppm C3H8
in air
Gas mixture contain-
ing (nominal) 50 ppm
hexane in air
Gas mixture contain-
ing (nominal)
20 ppm toluene
Gas mixture contain-
ing (nominal) 100
ppm mehanol in air
Check specifications,
conduct calibration
check which should be
within 52 of initial
NMO response factor
As shown above
As shown above
As shown above
Return to
manufacturer,
check new gas
As shown above
As shown above
As shown above
-------
o
o
o
-------
Section No. 3.17.2
Date May 31, 1991
Page 1
2.0 CALIBRATION OF APPARATUS
Calibration of the apparatus is one of the most important functions in
maintaining data quality. The detailed calibration procedures included in this
section were designed for the sampling equipment specified in Method 25 and de-
scribed in the previous section. The calibration of the analytical equipment is
described in the section detailing the analytical procedures, Section 3«17-5.
Table 2.1 at the end of this section summarizes the quality assurance functions for
the calibrations addressed in this section. All calibrations including the analyt-
ical equipment should be recorded on standardized forms and retained in a calibrat-
ion log book.
2.1 Sample Metering and Volume Systems
2.1.1 Sample Tank Volume - The volume of the gas sampling tanks used to for
sampling and as intermediate collection vessels must be determined as follows:
1. Mark or number each tank so that it is uniquely identified.
2. Weigh each tank empty to the nearest 5 g-
3. Fill the tank with distilled or deionized water and reweigh to nearest 5
g-
4. Record the data on Figure 2.1 or similar form and calculate the sample
tank volume.
5. Alternatively, the volume of the water used to fill each tank may be
determined to the nearest 5 nil. It is extremely important that all the
water is removed from the tank after calibration.
2.1.2 Volume of Sampling Train from Probe Tip to Sample Tank Valve - The volume of
the sampling train from the probe tip to the sample tank valve must be determined
as follows: '
1. Assemble the sampling train in the same manner as it will be used in the
field.
2. Leak check the sampling system as described below in Section 2.1.3-
3. After the train passes the leak check, attach a sample tank that has been
calibrated and is full of air at ambient pressure to the sampling system.
4. With the inlet still plugged from the leak check, open the sample tank
valve, flow control valve, sample valve, and purge valve to ensure that
the tank is at ambient pressure. Note that the mercury U-tube manometer
is reading zero.
5. Shut the sample tank valve and start the purge pump (or other pump of the
tester's choice).
6. After approximately 2 min, shut the purge valve and then turn off the
purge pump.
7. Open the sample tank valve and read and record the vacuum.
8. Record these data and the barometric pressure on Figure 2.2 or similar
form and then calculate the volume of the sampling train.
9. Repeat steps 4 through 8 twice. The calculated sampling train volume used
for the leak checks during testing will be the average of the three
calculated volumes. If the equipment is similar or the components are
interchangeable for all the Method 25 sampling systems used, the volume
from the probe tip to the sample tank valve only needs to be determined
for one system.
-------
Section No. 3-17-2
Date May 31, 1991
Page 2
o
Sample
Tank No.
Date
Calibra-
ted
Initial
Weight/Volume ,
g/ml*
Final
Weight/Volume ,
g/ml*
...
Sample Tank
Volume ,
ml*
Calibrated
By,
initials
••'••:
• •
*Weight measured to the nearest 5 £ or volume measured to the nearest 5 ol.
Figure 2.1. Sample tank and intermediate collection vessel volume determination.
O
o
-------
Section No. 3.17.2
Date May 31, 1991
Page 3
Calibrated By
Date Calibrated
System Assembled
Tank No.
System Leak Checked
Tank Volume, ml
Barometric Press., mm Hg
Are sampling train components similar and/or interchangeable?
Will the calculated sampling train volume be used for all trains?
Run
No.
1
2
3
Vacuum
Reading, mm Hg
Calculated
Volume , ml
Remarks
Average calculated volume
of sampling system
ml
Calculated volume, ml =
Calculated volume, ml -
Vacuum reading, mm Hg X Tank volume, ml
Barometric press, mm Hg - Vacuum reading, mm Hg
( ) X ( ) = ml
Average volume, nil
= Run 1 + Run 2 + Run 3 =
3
ml
Figure 2.2. Determination of sample train volume from probe tip to sample valve.
-------
Section No. 3.17.2
Date May 31, 1991
Page 4
2.1.3 Rotametevs - Two rotameters are needed, one for purging the sampling
system and a second for controlling the rate of sample collection. Since the
sampling rotameter is used to determine flow rate and maintain a constant
sampling rate, it must have an accuracy of +_ 10% for the flow rate used (60 to
100 cc/min) and a precision of +_ 5% over its range. The rotameter used to purge
the sampling train serves only as an indicator of the flow rate and its readings
are not used in any emission calculations; therefore, the accuracy of +_ 10% and
precision of +_ 5# do not apply after its initial calibration.
Initial Calibration - Both rotameters should be calibrated as part of the
sampling system when first purchased' and at any time the posttest check yields a
calculated sample volume that is not within 10% of the actual sample volume (for
sampling rotameter only) or erratic behavior is noted (for both rotameters). A
calibrated wet test meter, calibrated dry gas meter, or a properly sized bubble
meter should be used to calibrate the rotameters.
Before its initial use in the field, each rotameter should be calibrated as
part of the entire sampling system as described below.
1. Leak check the rotameter as part of the sampling system as follows:
a. Temporarily attach a suitable rotameter (e.g., 0-40 cm3/min) to the
outlet of the purge system.
b. Plug the inlet to the probe. Shut the sample tank valve and open the
flow control valve, purge valve, and the sample valve. Evacuate the
entire sampling system to 10 mm Hg.
c. Note the flow rate as indicated by the rotameter.
d. A leak of £0.02 L/min must be obtained; leaks >0.02 L/min must be
eliminated.
e. Close the purge valve and turn off the pump.
f. Note the vacuum reading.
g. Wait five minutes and take another vacuum reading.
h. If the pressure has changed by more than 20 mm Hg, the leak should be
found and corrected.
2. Attach a wet test meter, bubble meter, or calibrated dry gas meter to
the inlet of the probe.
3- Run the pump for 15 minutes with the flow rate set in the midrange (80
cc/min) to allow the pump to warm up and to permit the interior surface
of the wet test meter to become wet.
4. Collect the information required in the forms provided [Figure 2.3A
(English units) or Figure 2.3B (metric units)] using sample volumes
equivalent to at least five revolutions of the test meter or 10 minutes,
whichever is greater. Three independent runs must be made covering the
top, middle, and bottom of the flow rate range (i.e., 60, 80, and 100
cc/min).
5. Calculate the Yj for each run as shown on the data forms. The Y should
be in the range of 0.9 to 1.1 and the values of Yt should be in the
interval Y +_ 0.05Y, where Y is the average for three runs. If not, re-
calibrate, repair, or replace the rotameter. Otherwise, the Y (calibra-
tion factor) is acceptable and is to be used for future checks and
subsequent test runs. Alternatively, if the Yt's are acceptable and the
Y is outside the range of 0.9 to 1.1, the rotameter may be remarked to
reflect the corrected readings. The corrected readings will then be
used for testing and future calibrations. The completed form should be
forwarded to the supervisor for approval, and then filed in the calibra-
tion log book.
o
o
o
-------
Date
Calibrated by
Meter system no.
Primary meter no.
Barometric pressure, Pn = in. Hg Ambient temperature
, dry gas , or bubble meter
Type of primary meter: wet test
Primary meter readings
Initial
reading
(vpl),a
ft3
Final
reading
(vpf),a
ft3
Initial
temp,°F
(tpl)
op
Final
temp,°F
(tpf)
op
Press
drop
(Vc
in.
W
Rotameter readings
Initial
reading
(vs,).b
ft3 or
ft3 /min
Final
reading
(vs,)b
ft3 or
ft3 /min
Initial
temp
op
Final
temp
(tsf)
op
Press
drop
(Ds),c
in.
H20
Time
min
(e),d
min
Calibration
factors
(YJ,"
(Y)
0 Volume passing through the meter using the initial and final readings; requires a minimum of at
least five revolutions of the meter.
b Volume passing through the meter using the initial and final readings or the indicated flow rate
using the initial and final flow rate setting.
c Pressure drop through the meter used to calculate the meter pressure.
d The time it takes to complete the calibration run.
e With Y defined as the average ratio of volumes for the primary meter compared to the flowmeter
calibrated, Y must be 0.9 to 1.1 and Y1 = Y + 0.05Y thus:
*i '
F] [P. * (D/13.6)]
Vtl)/2je[(tpl + tpf)/2 + 460°F][Pm
-(Eq. 2-3),Y =
Y + Y •*• Y
xl *2 I3
-(Eq.2-4)
Figure 2.3A. Rotameter calibration data form (English units).
T3 O CO
{0 P (D
oq ct o
CD (D rt
M-
3 O
U1 P 3
OO O
LO
VD
to
-------
Date
Calibrated by
Barometric pressure, Pra =
Type of primary meter: wet test
Meter system no.
mm Hg Ambient temperature
Primary meter no.
°C
, dry gas
, or bubble meter
Primary meter readings
Initial
reading
(vpl),a
m3
Final
reading
b
m* or
m3 /min
Final
reading
(V)b
m3 or
m3 /min
Initial
temp
°C
Final
temp
(tsf)
°C
Press
drop
(Ds).c
mm
H20
Time
min
(9)/
min
Calibration
factors
(YJ."
(Y)
0 Volume passing through the meter using the initial and final readings and requires a minimum of at
least five revolutions of the meter.
b Volume passing through the meter using the initial and final readings or the indicated flow rate
using the initial and final flow rate setting.
0 Pressure drop through the meter used to calculate the meter pressure.
d The time it takes to complete the calibration run.
e With Y defined as the average ratio of volumes for the primary meter compared to the rotaineter
calibrated, Y must be 0.9 to 1.1 and Yj = Y +_ 0.05Y thus,
(Vpf -'VpilKt.i * fcsf)/2 + 273°K][Pn * (Dp/l3.6)]
r + Vsl)/2]9[(tpl * tpf)/2 + 273°K][PB + (E
(Eq. 2-7), Y =
(Eq. 2-8)
Figure 2.3B. Rotameter calibration data form (metric units).
13 o en
P f» CO
m ct o
(D CO ff
H-
s: o
o o
2:
cu o
LO
O
o
o
-------
Section No. 3.17.2
Date May 31. 1991
Page 7
Posttest Calibration Check - After each field test series, calculate the
volume of sample that should have been collected during the sample run and
compare the results with the actual volume collected. If the calculated volume
and the actual volume collected are within 10% of each other on the average for
the three runs, no recalibration is needed. If the average sample volumes are
not within 10%, the sampling rotameter should be recalibrated. When either the
sampling rotameter or purge rotameter exhibits erratic behavior during sampling
or purging of the system, it should be recalibrated. Performance of a posttest
calibration does not necessitate changes in the emission calculations.
2.3 Thermocouples
The thermocouples on the sample probes and the filter heating system should
be initially compared with a mercury-in-glass thermometer that meets ASTM E-l No.
63C or 63F specifications:
1. Place the thermocouples to be calibrated and the mercury-in-glass
thermometer in a bath of boiling water. Compare the readings after
the bath temperature stabilizes and then record them on the calibration
data form, Figure 2.4 or equivalent.
2. Allow both the thermocouple and reference thermometer to come to room
temperature. Compare the readings after the temperature readouts
stabilize.
3. The thermocouple is acceptable if the values agree within 3°C (5.4°F) at
both points.
4. Prior to each field trip, compare the temperature reading of the
mercury-in-glass thermometer at room temperature with that of the
thermocouple that is part of the metering system. If the values are not
within 3°C (5.4°F) of each other, replace or recalibrate the
thermocouple.
5. No posttest calibration of the thermocouples is required unless they
demonstrated erratic behavior during the sampling.
2.*f Barometer
The field barometer should be adjusted initially and before each test series
to agree within 2.5*1 mm (-0.1 in.) Hg with a mercury-in-glass barometer'or with
the pressure value reported from a nearby National Weather Service Station and
corrected for elevation. The tester should be aware that the National Weather
Service readings are normally corrected to sea level; uncorrected readings should
be obtained. The correction for the elevation difference between the weather
station and the sampling point should be applied at a rate of -2.5 mm Hg/30 m (-
0.1 in. Hg/100 ft) elevation increase, or vice versa for elevation decrease.
2.5 Absolute Pressure Gauge
The absolute pressure gauge should be calibrated against a mercury U-tube
manometer upon receipt and every quarter thereafter or upon erratic behavior.
Attach the the absolute pressure gauge and mercury U-tube manometer to a "T"
connection with a vacuum pump. Compare the readings at atmospheric pressure.
Pull a vacuum of 10 mm Hg of absolute pressure. Pressure readings should agree
within 3 B™ Hg. If this criteria is not met, make adjustments and repeat the
calibration.
-------
Date
Reference
thermometer
type
Calibi
thermc
type
-
'ated
jcoupl«
use
i
no.
Anbier
refera
it teinpei
calibrb
Heasurec
-ature
differc
I values
Bo:
refer8
.ling wai
calibrb
;er
differc
Calibrator's
initials
0
01
O
p
rt O
Temperature reading of the reference thermometer in °C or °F.
Temperature reading of the thermocouple being calibrated in °C or °F.
Difference between the reference thermometer and the calibrated thermocouple. This difference must
be less than 3°C (5.4°F) for than initial calibration and 6°C (10.4°F) for the calibration check.
Figure 2.4. Thermocouple calibration form.
O CD ft
3 O
000 3
CO O
CO
I-* •
VD H»
VD-J
I-' •
rvj
O
O
O
-------
Section No. 3.1?.2
Date May 31. 1991
Page 9
Table 2.1. ACTIVITY MATRIX FOR CALIBRATION OF EQUIPMENT
Apparatus
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Sample tank volume
Within 5 g or 5 ml
Calibrate initially
and when not within
10% of calculated
volume or shows
erratic behavior
Repair or
replace and
recalibrate
Sampling train
volume
No limits
Calibrate initially
as described in
Section 2.1.2
Repeat
Rotameters
Y = 0.9 to 1.1 and
all Y = Y +_ 0.5 Y
Calibrated initially
and when calculated
volume not within 10%
or erratic behavior
Repair, replace,
and recalibrate
Thermometers
Within 3°C {5.
of true value
Calibrate initially
as a separate com-
ponent with mercury-
in-glass thermometer;
check before each
test against mercury-
in-glass thermometer
Adjust or replace
Barometer
Within 2.5 mm
(0.1 in.) Hg of
mercury-in-glass
barometer or weather
station value
Calibrate initially
using mercury-in-
glass barometer;
check before and
after each test
Adjust to
agree with
certified
barometer
Absolute
Pressure
Gauge
Within 3 mm Hg of
mercury U-tube
manometer
Calibrate initially
using mercury U-tube
manometer; repeat
every quarter or
upon erratic behavior
Adjust to
agree with
mercury
U-tube
manometer
-------
o
o
o
-------
Section No. 3.17-3
Date May 31, 1991
Page 1
3.0 PRESAMPLING OPERATIONS
This section addresses the preparation and packing of supplies and equipment
needed for the sampling. The pretest sampling checks (Figure 3-1) is a listing of
equipment pretest operational checks which should be completed before leaving for
the field. The pretest preparations form (Figure 3-2) can be used as an equipment
packing checklist. The quality assurance activities for the presampling operations
are summarized in Table 3-1 at the end of this section.
3.1 Preliminary Survey
The preliminary survey may be needed to properly design the final emission
sampling and analytical protocol. Preliminary measurements may include
determining the stack dimensions, the flue gas moisture, stack pressure, and stack
temperature, if the concentration of organics is to be determined on the basis of
stack conditions. Other measurements which may be made, depending upon the
requirements of the applicable regulation and the source operations, include a flow
rate determination, velocity check, and stack gas temperature range. The tester
must determine these requirements and use the proper test methods to establish
these data (i.e., EPA Methods 1 through 4).
One of the primary concerns for any organic sampling program must be safety.
The tester should always question the facility representative concerning general
plant safety requirements and safety in regard to sampling at the selected sampling
site. Every sampling and analysis protocol should address the safety
considerations involved in performing the protocol. Because there are numerous
safety considerations involved in organic sampling, it is beyond the scope of this
Handbook to discuss each one in detail. However, it cannot be over-emphasized that
the tester must always be aware of the safety hazards.
Figure 3-3 may be used to collect preliminary survey data.
3.2 Checking and Calibrating the Apparatus
3.2.1 Sampling System - The Method 25 sampling train (see Figure 1.1) is
commercially available or can be manufactured in-house if the apparatus complies
with specifications in the EPA Test Method (see Section 3.17.10). These
specifications and other performance criteria are discussed in greater detail in
Section 3•17•1•
Heated Probe - The probe's thermocouple should have been initially calibrated
as described in Subsection 3.17.2.3. Prior to each field test, the thermocouple
should be attached to the temperature readout device and the probe thermocouple
reading should be compared with a reference thermometer reading at the ambient
temperature. If the values are not within 3°C (5.4°F) of each other, replace or
recalibrate the thermocouple.
The probe should be cleaned with acetone or methylene.chloride, nitric acid,
and distilled or deionized water. To dry the probe, turn on the probe heater and
purge the probe with UPC air or nitrogen. The objective is to leave the probe free
of contamination.
Leak check the probe and check the probe heater system as part of the complete
train assembly as described below.
-------
Section No. 3.17-3
Date May 31, 1991
Page 2
Date Calibrated by
Heated Probe
Heater system capable of maintaining heat at probe exit at 129°C? yes no
Thermocouple calibrated against reference thermometer? yes no (within 3°C)
Probe cleaned with soap and water and acetone, then dried? yes no
Probe leak checked with sampling system? yes no
Filter Holder/Heating System
Heater system capable of maintaining heat at filter exit at 121°C? yes no
Thermocouple calibrated against reference thermometer? yes no (within 3°C)
Assembly cleaned with soap and water and acetone, then dried? yes no
Filter assembly leak checked with sampling system? yes no
Condensate Trap
Traps burned at 300°C until acceptable blank level achieved? yes no
Nitrogen placed in trap and trap sealed? yes no
Purge System
Rotameter initially calibrated? yes no (accuracy lOJt, precision 5%)
Purge system checked with sampling system and indicated flow? yes no
Sampling Rotameter
Initially calibrated against primary standard? yes no (accuracy 10%,
precision 5%)
Calculated sample volume within 10% of actual sample for last test?
yes no
Sample Tank
Tank volume calibrated gravimetrically or volumetrically? yes no (+_ 5 ml)
Tank cleaned and filled with nitrogen? yes no
Barometer or Absolute Pressure Gauge
Calibrated against mercury-in-glass barometer? yes no
(l 2.54 mm Hg)
Ambient Thermometer
Initially calibrated against reference thermometer? yes no (+_ 1°C)
Figure 3-1- Pretest sampling checks.
o
o
o
-------
Section No. 3.17-3
Date May 31, 1991
Page 3
Apparatus check
Sampling System
Heated Probe
Heating checked
Thermocouple
calibrated*
Cleaned*
Leak checked
Filter Holder/Heater
Heater checked
Thermocouple
calibrated*
Cleaned*
Leak checked
Condensate Trap
Burned & blank
checked*
Sealed*
Purge System
Pulls flow
Sampling Eotameter
Calibrated or checked*
Sample Tanks
Calibrated*
Cleaned*
Pressure Gauge
Calibrated*
Ambient Thermometer
Calibrated or checked
Sampling Supplies
Dry ice
Fi Iters
Free of organic
binder*
Acceptable
Yes
No
Quantity
Required
Ready
Yes
No
Loaded and Packed
Yes
No
*Most significant items/parameters to be checked.
Figure 3^2. Pretest preparations,
-------
Section No. 3.17-3
Date May 31. 1991
Page 4
I. Name of company
Address
Contacts
Date_
o
Phone
Process to be sampled_
Duct or vent to be sampled_
II. Process description_
Raw material
Products
Operating cycle
Check: Batch
Continuous
III.
Timing of batch or cycle
Best time to test
Sampling site
. Description
Site description
Cyclic_
Duct shape and size_
Material
O
Wall thickness
Upstream distance
Downstream distance
inches
inches
inches
diameter
diameter
Size of port
Size of access
Hazards
area
B.
Ambient temp
Properties of gas stream
Temperature _ _°C
Velocity
F,
Static pressure
Moisture content
Particulate content
Gaseous components
Data source
, Data source
inches H20, Data source
%, Data source
, Data source
N2
02
CO
C02
S0
% Hydrocarbons (ppm) Toxics/Acids (ppm)
H2S
HC1
HF
Other
(Continued)
Figure 3-3- Preliminary survey data sheet.
O
-------
Section No. 3.17-3
Date May 31, 1991
Page 5
Figure 3-3 (Continued)
Hydrocarbon components
PPm
ppm
ppm
ppm
ppm
ppm
C. Sampling considerations
Location to set up sample recovery area
Special hazards to be considered_
Power available at duct_
Power available for GC
Plant safety requirements_
Vehicle traffic rules
Plant entry requirements
Security agreements
Potential problems
Safety equipment (glasses, hard hats, shoes, etc.)
D. Site diagrams. (Attach additional sheets if required).
IV. On-site collection of preliminary survey samples
A. Evacuated tank
Tank have been cleaned, heated in furnace and purged
with nitrogen?
Tank evacuated to the capacity of pump?
Filter end of probe placed at center of stack, probe
purged and sampled collected into flask until flask is at
stack pressure?
Stopcocks closed and taped?
Duct temperature and pressure recorded?
B.
Purged flasks
Flasks cleaned and purged with nitrogen?
Filter end of probe placed into stack, sample purged for
2 to 5 min and then stopcocks closed?
Stopcocks taped to prevent leakage?
Duct temperature and pressure recorded?
Stability and adsorption checks conducted?
(Continued)
-------
Section No. 3.17-3
Date May 31, 1991 ^-s.
Page 6 ( )
Figure 3.3 (Continued)
C. Quality assurance performance audit samples
Range of emissions to order proper range of performance
audit samples inlet ppm outlet ppm
Address to send audit samples
D. Bulk samples and screening techniques
Bulk emission sample(s) collected?
Bulk liquid sample(s) collected?
Detector tubes or other screening techniques used?_
E. Safety with respect to sample collection
Can gases be exhausted through purge system? yes no
Can electrical service be used? yes no
Can heated probe and heated filter be used? yes no
Can thermocouple be used? yes no
F. Emission results must be reported in terms of:
ppmC at standard conditions
ppmC at stack conditions
2
ppmC at standard conditions corrected for CO
ppmv as a related solvent at standard conditions
ppmv as a related solvent at stack conditions
pounds per hour of carbon '
o
pounds per hour of related solvent
O
-------
Section No. 3-17-3
Date May 31, 1991
Page 7
Filter Holder and Heating System - The filter holder assembly's thermocouple
should have been initially calibrated as described in Subsection 3-1?.2.3. Prior
to each field test, the thermocouple should be attached to the readout and the
probe thermocouple reading should be compared with a reference thermometer reading
at the ambient temperature. If the values are not within 3°C (5.4°F) of each
other, replace or recalibrate the thermocouple.
The filter holder/heating assembly should be cleaned with acetone or methylene
chloride, nitric acid, and distilled or deionized water. To dry the filter holder,
turn on the probe heater and purge the filter holder with UPC air or nitrogen. The
objective is to leave the assembly free of contamination.
Leak check the filter holder assembly and check the filter heater system as
part of the complete train assembly as described below.
Condensate Trap - Before its initial use and after each subsequent use, a
condensate trap should be thoroughly cleaned and checked to insure that it is not
contaminated. Both cleaning and checking can be accomplished by installing the
trap in the condensate recovery system and heating to 300°C while performing a
system background test (described in Section 3«17-5-2.1). A trap may be considered
clean when its effluent concentration is below 10 ppm. Clean or "blanked" traps
should be filled with nitrogen and sealed to prevent contamination and corrosion.
If a trap cannot be properly "blanked," it should be discarded. The history of the
trap should also be tracked. It is recommended that traps previously used for
inlet sampling be used only for inlet sampling thereafter.
Purge System - The purge system must be capable of purging the probe and
filter holder assembly at a rate of 60 to 100 cc/min. Upon initial receipt, the
rotameter should be calibrated as described in Section 3-17-2. After the initial
calibration, the rotameter and pump should be checked prior to each field test to
ensure that it is capable of purging the sampling system at the rate indicated
above. The rotameter for the purge pump is used only as a flow indicator during
the field test.
To check the operations of the sampling system, the probe, filter
holder/heating assembly, and purge system should be assembled in the same manner
that it will be used in the field with the exception that (1) no filter is needed
and (2) the tank will not be attached. Check the system as follows:
1. Turn on the probe and the filter holder heaters.
2. With the inlet to the probe open, turn the sample tank valve off, turn the
purge pump on, set the flow rate to about 80 cc/min, and allow air to be
drawn through the system until the operating temperatures are met.
3. After the probe and filter holder comes to the proper temperatures (probe
129°C and filter 121°C), plug the inlet to the probe and conduct a leak
check as described in Subsection 1.1.10. If the system does not pass the
leak check, repair or replace the faulty component(s) and repeat the leak
check until it is acceptable.
Sampling Eotameter - The sampling rotameter must be calibrated (1) initially
as described in Subsection 2.1.3, (2) when the posttest calculated volume from the
previous field test is not within 10% of the actual sample volume collected and
(3) any time the rotameter exhibits erratic behavior. If the rotameter is
acceptable using the criteria above, the tester may attach the purge pump after the
sample tank valve and ensure that the rotameter does not exhibit erratic behavior
by pulling a flow of about 80 cc/min. If erratic behavior is exhibited, the
rotameter should be cleaned and recalibrated.
•it'
-------
Section No. 3-17.3
Date May 31. 1991 ^-^
Page 8 f )
Sample Tanks - Each sample tank must be initially calibrated as described in
Section 3.17.2. After the initial calibration, it should be visually checked prior
to each field test to ensure that there are no dents that would effect the total
sample volume. Each tank must also be flushed with UPC air until there is no
response on the NMO analyzer. After a tank is free of hydrocarbon contamination,
it should be evacuated and filled with nitrogen to a pressure of approximately 10
mm Hg above atmosphere. The nitrogen will prevent contamination and corrosion.
Mercury Manometer or Absolute Pressure Gauge - If a mercury manometer is used,
then a barometer will be required. No calibration is required for a mercury
manometer. The manometer or absolute pressure gauge should be leak checked as part
of the assembly as described above. The barometer and/or absolute pressure gauge
should be calibrated against a mercury-in-glass barometer as described in
Subsection 3.17.2.4. If it does not agree within 2.54 mm Hg, it should be
corrected or replaced.
Thermometer for Ambient Temperature - A thermometer is needed to measure the
ambient temperature. If the thermometer is a mercury-in-glass thermometer with a
sensitivity of 1°F, the thermometer should be initially check against a mercury-in-
glass ASTM thermometer as described in Section 3-17.2. After the initial
calibration, the thermometer should only be visually checked to ensure that it is
not broken. If a thermocouple is used to measure the ambient temperature, it
should be checked at ambient temperature against a reference thermometer and should
agree with 1°C.
3.2.2 Sampling Supplies - The following supplies are needed for sampling:
Crushed Dry Ice - Crushed dry ice is needed to cool the condensate trap during
sampling for better collection of organics and to keep it cool until analysis.
There is no specification on the dry ice.
Filters - Glass fiber filters, without organic binder are needed to remove
organic particulate matter from the gas stream during sample collection.
Typically, the filters used for EPA Method 5 sampling are satisfactory, if no
organic binders are present. If organic binders are present, they may be released
during testing and positively bias the results. If the tester is not certain about
the presence of organic binders in a glass fiber filter, it should be placed in a
furnace at 300°C for 2 hours to remove any organic binders present. This
procedure, however, may make the filter more brittle resulting in a greater need
for caution in handling. A check on the amount of organic binder lost during
heating can be determined by following the procedures described in Subsection
1.3.1.
3-3 Packing the Equipment for Shipment
The sampling system is relatively small, made predominately of stainless
steel, and therefore rugged with the exception of the rotameters, pumps, and
pressure gauges. The filter holder system and probe may be packed separately or as
a unit. The other components should be packed conveniently and securely in
labeled containers (as to contents) for ease of identification in the field.
Polyethylene foam can be used to cushion the components. Also, the tanks should be
secured so they do not become dented.
.1
o
o
-------
Section No. 3.17.3
Date May 31, 1991
Page 9
Table 3.1. ACTIVITY MATRIX FOR PRESAMPLING PREPARATION
Operation
Apparatus
Check and
Calibration
Heated probe
Filter holder
Condensate
trap
Purge system
Rotameter
Sample tank
Pressure
gauge
Sampling
Supplies
Dry ice
Filters
Packing Equip-
ment for Ship-
ment
Acceptance limits
Leak free, cleaned,
capable of heating
to 129° C with cali-
brated thermocouple
Leak free, cleaned,
capable of heating
to 121° C with cali-
brated thermocouple
Acceptable blank
level
Capable of purging
at rate of 80 cc/min
Calibrated
Volume calibrated
to +_ 5 ffll and
clean
Range 0 to 900 mm
Hg within 2.5^ mm Hg
No specifications
Glass fiber with
no organic binders
Packed in secure
container
Frequency and method
of measurement
Clean with soap and
water, then acetone;
calibrate thermocouple
against reference
thermometer; conduct
heater check and leak
check with assembly
As above
Check as described in
Subsection 3.17.5,
Check as part of
sampling assembly
Calibrate as shown
in Section 3.17.2
Calibrate and clean
as described in
Subsection 3.17.1.1
Check against mercury-
in-glass barometer
Not applicable
Check or heat filter
Before field trip,
pack in shipping
container
Action if require-
ments are not met
Repeat cleaning,
calibration, and/or
heater and leak
checks
As above
Repeat burnout and
repeat blank check
or replace and repeat
Repair and repeat
Clean and recalibrate
Recalibrate and/or
reclean
Adjust and repeat
calibration
Not applicable
Replace
Repack
-------
o
o
o
-------
Section No. S.I?.1
Date May 31. 1991
Page 1
4.0 ON-SITE MEASUREMENTS
On-site activities include transporting the equipment to the test site,
unpacking and assembling the equipment, sampling for total gaseous nonmethane
organics, and recording the data. The associated quality assurance activities are
summarized in Table 4.1 at the end of this section.
4.1 Transport of Equipment to the Sampling Site
The most efficient means of transporting the equipment from ground level to
the sampling site (often above ground level) should be decided during the
preliminary site visit or by prior correspondence. Care should be taken to
prevent damage to the equipment or injury to test personnel during the moving. A
clean "laboratory" type area free of excessive dust and organic compounds should
be located and designated for preparing the tanks, traps, and filter holder, and
for sample recovery.
4.2 Preliminary Measurements and Setup
It is recommended that a preliminary survey be conducted prior to sampling and
analysis, unless adequate prior knowledge of the source and/or information, is
available. Testing must be conducted at the proper sampling locations and during
the proper process and control equipment operating cycles or periods. The tester
should refer to Subsection 3«17«3«1 regarding the information that is typically
needed to establish the proper sampling and analysis protocol. The accuracy of
sampling and analysis following handling and transportation of the sampling system
to and from the sampling site is determined using a cylinder gas audit.
One of the primary concerns for any organic sampling program must be safety.
The tester should always question the facility representative concerning general
plant safety requirements and safety in regard to sampling at the selected sampling
site. Every sampling and analysis protocol should address the safety considera-
tions involved in performing the protocol. Because there are numerous safety
considerations involved in organic sampling, it is beyond the scope of this
Handbook to discuss each one in detail. However, it cannot be over-emphasized that
the tester must always be aware of the safety hazards.
4.3 Sampling
Because of the unlimited variations in sampling organic compounds from
potential source types, only the more general common situations and problems are
addressed in this section. Both required and recommended quality assurance/control
checks and procedures are provided to assist in the collection of acceptable
quality data and to assess the accuracy of the sampling and analysis.
On-site sampling includes the following steps:
1. Conducting preliminary measurements and setup of the recovery area.
2. Preparation and setup of the sampling system.
3. Connection of electrical service and leak check of the sampling system.
4. Heating the probe and filter to proper temperature.
5. Insertion of the probe into the duct and sealing the duct.
6. Purging of the sampling system.
7- Constant rate sampling.
8. Recording data.
/•
(-I
-------
Section No. 3.1?.4
Date May 31, 1991
Page 2
9. Recovering the sample and its transportation to the laboratory.
4.3.1 Preliminary Measurements and Setup of Recovery Area - All test personnel
should be knowledgeable of the plant safety requirements. This includes areas that
should not be entered and whether the plant will allow the use of electrical
service in the sample collection area. The risk to test personnel of being in
process areas handling organic compounds, removing gases from these processes, and
venting sample gases into the sample collection area may be significant. The plant
may require that no electrical service be used and/or that the sample gases removed
from the process during the pretest purging of the sampling system be adsorbed onto
some vehicle such as a resin or charcoal or vented back into the process.
After all testing personnel are familiar with plant safety requirements and
all potential safety hazards, preliminary measurements and setup may begin. The
sampling site should be checked to ensure that adequate electrical service is
available (if allowed). The stack dimensions are measured and recorded (if
applicable) on a field sampling data sheet such as the one shown if Figure 4.1.
If the concentration of the organics are to be determined on any basis other than
ppmC at standard conditions, the corresponding preliminary measurements should be
made at this time. Moisture content of the stack cannot be measured by Method 25.
Therefore, if the final emissions are to be presented in terms of stack
concentration or mass emission rate, or are to be corrected to a C02 concentration,
or to an equivalent solvent basis, Methods 1 through 4 will most likely need to be
conducted simultaneously with Method 25 • Prior to final sampling, the tester must
determine on what basis the final results are to presented.
If the emissions are to be presented in terms of a mass emission rate, the
flow rate of the stack gas, including its moisture content, must be determined. In
this case, it is preferable that the sampling location be selected in accordance
with Method 1 (or Method 1A, if applicable). If this is impractical, it should be
selected to minimize flow disturbances. The number and locations of sampling
points for the velocity traverse are selected according to Method 1 (see Section
3.0:1 of this Handbook); the traverse is conducted according to Method 2 (see
Section 3.1 of this Handbook). Note; The Method 25 sampling is conducted at a
single point of average velocity. If it is unsafe to conduct a preliminary
velocity traverse or a traverse is not required, the sampling may be conducted at
the center of the duct or at a point at least 3 feet into the duct (whichever is
less). The port must be sealed well and there must be no reason to suspect that
the emission concentration is not uniform across the stack.
Method 25 requires constant rate sampling; the sampling rate is not changed
with regard to the flue gas flow rate. However, if the emissions are presented on
a mass emission basis, the flue gas flow rate must be measured during each Method
25 sample run and the corresponding flow rate used to determine the mass emission
rate for that run.
Select a total sampling time greater than or equal to the minimum sampling
time specified in the applicable subpart of the regulation or other applicable
emission standard. The data will be recorded at 5-minute intervals.
A clean "laboratory" type area should be found to load the filter, recover the
samples, conduct orsat analyses, prepare and recover the moisture sampling train,
and to store other sampling equipment. This area should be free of excess dust or
high levels of organics. Because of the relatively small size of the Method 25
sampling equipment and the nature of Method 25 sample recovery, the stack location
can often be used as the recovery area.
o
o
o
.(i
-------
Section No. 3.17-4
Date May 31. 1991
Page 3
Company Name
Address
Run No.
Date
Sampling Location ; Start Time _
Tank No. Trap No. Sampling Train No. Finish Time
Thermocouple No. Barometer No. Operator
Vacuum Gauge No.
Tank Volume liters Train Volume
cc
Calculated Allowable Change p
cm Hg
TANK PARAMETERS
Parameter
Pretest
Posttest
Barometric
Pressure
in. Hg
mm Hg
Tank
Temp.
op
°C
Final Pressure
Tank Pressure
mm Hg
Gauge
Absolute
Leak Check
(cm Hg/min)
Tank
System
SAMPLE DATA
Clock Time
Tank Vacuum or
Gauge Pressure,
in. Hg (mm Hg)
Flow Meter
Setting, cc/min
Sample Gas Temp., °F (°C)
Probe Exit
Filter Inlet
Figure U.I. Field sampling data form.
-------
Section No. 3.1?.4
Date Hay 31, 1991 ^-^
Page 4 f J
4.3.2 Sampling System Preparation - Sampling system preparation includes (1) leak
checking the tanks, (2) assembling the sampling train, and (3) leak checking the
entire sampling system. This preparation should be conducted as follows:
1. The sample tanks should be evaluated to 10 mm Hg or less absolute
pressure. The pressure must be measured with a mercury U-tube manometer
or absolute pressure gauge capable of measurement within 1 mm Hg. Record
the tank pressure on the field sampling data form (Figure 4.1 or similar
form). Close the sample tank valve and allow the tank to sit for 60
minutes. The tank pressure should then be rechecked using the same
pressure gauge and this pressure recorded. If after 60 minutes no
noticeable change (less than 4 mm Hg) has occurred in the tank pressure,
the tank is acceptable for testing. The tank evacuation and leak check
may be conducted either in the laboratory or the field.
2. Just before sampling train assembly, measure the tank vacuum using a
mercury U-tube manometer or absolute pressure gauge capable of measurement
to within 1 mm Hg. Record this pressure, the ambient temperature, and the
current barometric pressure on the field sampling data form.
3. Close the sample tank valve and assemble the sampling system as shown in
Figure 1.1.
4. Immerse the condensate trap body in dry ice. The point where the inlet
tube joins the trap body should be 2.5 to 5 cm above the top of the dry
ice.
5. After assembling the sampling train, plug the probe tip, and make certain
that the sample tank valve is closed. /"""N
6. Turn on the purge vacuum pump, and evacuate the sampling system from the ( j
probe tip to the sample tank valve to an absolute pressure of 10 mm Hg or ^—
less and record the pressure on the field sampling data form.
7. Close the purge valve, turn off the pump, wait 5 minutes, and recheck the
indicated pressure and record this reading. The change in the absolute
pressure of the tank during the 5-min period is the measured pressure
change (delta P). Note; A less sensitive pressure gauge, which is
standard on commercially available equipment and reads pressure to the
nearest 0.5 in. Hg, can be used for indication of the vacuum during
testing, but cannot be used for measuring the tank pressure and for the
leak check procedure. The tester must conduct the leak check by attaching
the mercury U-tube manometer or pressure gauge at the probe inlet.
The sample tank must remain closed. If opened, the extremely large volume
of the tank compared to the sample system makes detecting a small leak
extremely difficult.
8. Calculate the maximum allowable pressure change based on a leak rate of 1
percent of the sampling rate using the equation below. Record it on the
field sampling data form. This should be compared to the measured P
from Step 7 which must be less than or equal to the allowable P
calculated in Equation 4-1.
P = 0.01 FPb9 Eq. 4-1
vt
Where:
P = Allowable change, cm Hg,
F = Sampling flow rate, cc/min,
Pb = Barometric pressure, cm Hg,
8 = Leak check period, min, and
Vt = Sampling train volume from probe tip to tank volume, cc.
•J/
o
-------
Section No. 3.17.4
Date May 31, 1991
Page 5
9. If the measured pressure change exceeds the calculated limit, correct the
problem before beginning sampling.
10. Mark the probe such that when placed in the stack, the nozzle will be at a
point of average velocity. Alternatively, if the flow cannot be measured
or does not have to be measured, mark the probe such that the nozzle will
be placed at the center of the duct or a point at least three feet into
the duct (whichever is less).
11. Calculate the sampling rate for the test by dividing sampling time
(minutes) into 80% of the sample tank volume (cc). The resulting flow
rate, in cc/min, will be used for the testing and will provide a margin of
error.
12. Complete all remaining entries on the field sampling data form that can be
completed prior to the start of the test run. This includes all facility,
sampling train component, and run information plus the barometric
pressure.
4.3.3 Constant Rate Sampling - Sampling must be conducted at a constant rate (+_
10%) over the duration of the sampling period at a rate of between 60 and 100
cc/min. Follow the steps below to obtain a constant rate sample:
1. Unplug the probe tip, and place the probe into the stack such that the
probe is perpendicular to the duct or stack axis and the probe tip (or
nozzle) is located at a point of average velocity with the tip facing away
opposite of the direction of gas flow. For ducts with a negative static
pressure, sufficiently seal the sampling port to prevent air in-leakage
around the probe.
2. Set the probe temperature controller to 129°C (265°F) and the filter
temperature controller to 121°C (250°F). Allow the probe and filter to
heat for about 30 minutes prior to purging the train.
3. To purge the probe and filter assembly, close the sample valve, open the
purge valve, and start the vacuum pump. Set the flow rate between 60 and
100 cc/min, and purge the train with sample gas for at least 10 minutes.
4. When the temperatures at the exit ends of the probe and filter are within
their specified ranges, sampling may begin.
5. Check the dry ice level around the condensate trap, and add dry ice, if
necessary.
6. To begin sampling, close the purge valve and stop the pump. Record the
start time. Open the sample valve and the sample tank valve. Using the
flow control valve, set the flow through the sampling train to the
calculated flow rate. Record the tank vacuum, flow rate setting, probe
temperature, and filter temperature.
7. Adjust the flow rate as necessary to maintain a constant rate •(+_ 10%)
throughout the sampling period. Also, adjust the probe and temperature
controllers as necessary to maintain the proper temperatures.
8. Record the sample tank vacuum, flowmeter settings, probe temperature, and
filter temperature at 5~°inute intervals.
9- Sample collection ends when the total sampling time is reached or when the
constant flow rate can no longer be maintained due to reduced sample tank
vacuum. If sampling must be interrupted before reaching the minimum
sampling time (specified in the applicable regulation) because a constant
flow rate cannot be maintained, proceed as follows: Record the sample
tank pressure and temperature, close the sample tank valve, remove the
used sample tank from the sampling train (without disconnecting other
parts of the sampling train). Take another evacuated and leak-checked
-------
Section No. 3.17-4
Date May 31, 1991
Page 6 /"""N
sample tank, measure and record the new tank vacuum, and attach the new
tank to the sampling train. After the new tank is attached to the
sampling train, proceed with sampling until the required minimum sampling
time has been reached.
10. After sampling is completed, close the flow control valve, and record the
final tank vacuum; then record the tank temperature and the barometric
pressure. Close the sample tank valve.
The sampling train will be recovered as described below. After the trap and
tank are disconnected from the sampling train, both the Sampling System Preparation
procedure (steps 1 through 12) and Constant Rate Sampling procedure (steps 1
through 10) are repeated for the next sample run. If the tester feels that the
filter will not become plugged during the subsequent sample run, the probe and
filter holder/filter assembly including the used filter may be used for any
proceeding sample runs.
A cylinder gas performance audit shall be conducted during the sample
collection phase of the test. The procedures for collecting the cylinder gas are
described in Section 3-17«8« Method 25 requires that each test be audited with two
concentrations of cylinder gases. Since the tester must have two sampling trains,
a regulator, connecting tubing, and a sampling manifold available for the audit,
these items must be included in the test protocol. The collection of the audit
gases should be conducted in the same manner as collection of the field samples
with the exception that the probe, filter holder, and purge system are not used for
collection of the audit gases. If the agency representative is not present at the
start of field sampling, the tester should wait until the conclusion of the field
test to conduct the audit; it is always preferable to have the
representative present during the audit. The tester should not break the seal
the audit cylinders until just prior to collecting the audit samples to provide the
agency with the maximum opportunity to observe all steps in the collection of the
field samples and the audit samples.
4.3.4 Sample Recovery - Prior to sample recovery, the flow control valve and the
sample tank valve should be closed and the field sampling data form should be
completed. If the sampling location is not suitable for conducting sample
recovery, the sampling train should be removed to the sample recovery area;
otherwise, the sample may be recovered at the sampling location. Samples should
be recovered as follows:
1. Disconnect the sample tank from the sampling system. If the less
sensitive pressure gauge (see Note in Subsection 4.3.2) was used for
sampling, the tank should be immediately attached to the more sensitive
gauge (reading to within 1 mm Hg) and the tank pressure recorded.
2. Disconnect the condensate trap at the flowmetering system, and tightly
seal both ends of the trap. The probe (from the stack to the filter) is
not included as part of the condensate sample.
3- Pack the trap in dry ice during storage and shipping and until the samples
are analyzed.
4. Ensure that the condensate trap and the sample tank(s) are properly
identified by the test run number and their corresponding identification
numbers are properly entered on the field sampling data form. The use of
a standardized label is encouraged and is helpful in ensuring consistent
identification by the laboratory staff. x—N
5. Label a clean condensate trap and tank as sample blanks. These will be( )
analyzed in the same manner as the field samples and then used tox J
determine the blank level of the sampling system.
-------
Section No. 3.17-4
Date May 31, 1991
Page 7
4.4 Sampling Problems
Because of the large number and variety of organic processes, it is not
possible to discuss all the sampling problems related to Method 25 sampling. Only
the seven most common problems will be addressed: (1) high sample gas moisture
content and freezing of the trap, (2) no use of electrical service in sampling
area, (3) conversion of C02 to a carbonate in the trap, (4) use of Method 25 for
measuring low levels of organics, (5) sample contamination, (6) sampling and
analysis by different companies, and (7) measurement in ducts containing organic
droplets.
4.4.1 High Gas Sample Moisture Content and Freezing of Trap - If the sample gas
has a high moisture content, the small line running from the filter to the
condensate trap tends to freeze and the moisture blocks the sample gas flow. For
this reason, the trap should clear the dry ice by 2.5 to 5 cm. If freezing of the
moisture in the line continues to be a problem, the line may be insulated to
improve heat transfer. If the problem persists and sampling cannot be conducted,
the tester may place a "preliminary" trap in front of the "primary" trap. The
"primary" trap should be placed in an ice bath and the second trap placed in the
dry ice bath. After sample recovery, both traps are placed in dry ice until
analysis.
4.4.2 Use Electrical Service Not Permitted for Probe and Filter Heating - If, for
safety reasons, the plant cannot allow the use of electrical service at the
sampling site, sampling should be conducted using an in-stack filter. The filter
should consist of a stainless steel tube packed with quartz wool, similar to the
sampling filter in the original version of Method 25. The condensate trap is
connected directly to the in-stack filter, and the sampling system is not purged.
Recovery of the condensate trap does not include the in-stack filter. The
condensate trap must comply with the revised Method 25 specifications.
'...-'-. t i .
4.4.3 Conversion of C02 to Carbonate in Trap - It has been demonstrated that if
ammonia is present in the duct during testing, the C02 can be converted to
carbonate during testing. This conversion results in a high bias during the
analysis of the trap. If this problem occurs, consult the Administrator for an
alternative procedure.
4.4.4 Use of Method 25 for Measuring Low Levels of Organics - Method 25 was not
intended to measure organics at levels below 50 ppmC. However, if the tester has
no other options. Method 25 can be used under the following conditions: (1) extreme
caution must be used in preparing the traps and tanks and (2) two traps and two
tanks should be set aside as field blanks with the analytical results subtracted
from the field sample values. This approach will improve measurements at lower
level sources, although the precision and accuracy of the method will be poor.
4.4.5 Sample Contamination - Sample contamination is a major problem with Method
25 sampling and analysis. Precautions to prevent contamination are listed below:
1. Pretest preparation of the probe, filter holder, traps, and tanks cannot
be overemphasized. The probe and filter holder assembly must be cleaned
in the manner prescribed to eliminate organic materials. The traps must
be burned after analysis to remove any organics. The traps should be
filled with nitrogen under pressure and their history should be tracked.
It is recommended that no trap previously used for sampling at high
-------
Section No. 3-1?.4
Date May 31, 1991
Page 8
organic levels be used for sampling extremely low levels of organics.
2. All components of the train should be maintained such that the surfaces
are never exposed to organics (e.g., oil or other organic compounds),
particularly the quick-connects or fittings.
3. All other components, such as the tubing used to connect the audit
cylinders to the sampling manifold, must be free of organics.
4.4.6 Sampling and Analysts by Different Companies - Because of the small number
of laboratories that conduct Method 25 analysis, a large portion of the Method 25
sampling and analysis is conducted by two different companies (i.e., the sampling
company and the analytical laboratory). This creates problems in assigning
responsibility when audit sample results are not acceptable. If the sampling
company wants to check the consistency of the analytical results (especially for
low level organic sampling), the tester should obtain extra traps and cylinders
from the analytical laboratory. These clean traps and cylinders should not be
opened, marked as if they were a sample, and submitted for analysis with the field
samples.
4.4.7 Measurement in Ducts Containing Organic Droplets - If organic droplets exist
in the duct to be sampled, the Method 25 results can be greatly biased. The tester
should first try to find another location where the droplets do not exist. If this
is not possible, two filters may need to be placed in the system with both being
replaced after each sample run. The addition of an in-stack filter should help
collect organic droplets and will reduce the loading on the out-of-stack (second)
filter.
4.5 Sample Logistics and Packing Equipment
The sampling and sample recovery procedures are followed until the required
number of (1) runs are completed, (2) audit samples are collected, and (3) blank
samples are labeled. Log all data on the sample recovery and integrity data form,
Figure 4.2. At the conclusion of the test:
1. Check all traps and tanks for proper labeling (time, date, location, test
run number, and any other pertinent documentation). Be sure that blanks
have been set aside and labeled.
2. If possible, make a copy of the field data form(s) in case the originals
are lost.
3. Examine all tanks and traps for damage and ensure that the traps are
packed in a sufficient amount of dry ice for transport to the base
laboratory. Ensure that the containers are labeled properly for shipping
to prevent loss of samples or equipment.
4. Review the field sampling data form and any other completed data forms to
ensure that all data have been recorded and that all forms are present.
o
o
-------
Plant Name
Sample Recovery Person
Plant Location
RECOVERED SAMPLES
Run
No.
Sample Type
Inlet Outlet Audit Blank
Trap
No.
Placed on
Dry Ice
Tank
No.
Date
Recovered
Time
Recovered
Remarks
LABORATORY CUSTODY
Date of Laboratory Custody
All traps still on dry ice?
Remarks
Laboratory Person Taking Custody
All samples identifiable?
Figure ^.2. Sample recovery and integrity data form.
LO C
H* •
*
«J
H* •
vo-
H1 •
J
-------
Section No. 3-17.4
Date May 31, 1991
Page 10
Table 4.1. ACTIVITY MATRIX FOR ON-SITE MEASUREMENTS
o
Characteristic
Preliminary
determinations
and
measurements
Sampling
system pre-
paration
(Continued)
Acceptance limits
If final results on
stack concentration
basis, determine the
moisture content of
stack gas
If final results on
emission rate basis,
determine moisture
content and flow
rate of stack gas
Determine stack
dimensions
Select sampling
time y minimum ,
total 'sampling time
in applicable!
emission standard;
number of minutes
between readings
shouldjbe an integer
Leak check tanks,
measure pressure
with manometer or
absolute pressure
gauge to within
1 mm Hg :
Assemble sample
train as shown in
Figure 1.1
Mark the probe such
that nozzle will
be at the point of
average velocity; if
flow can not or
does not need to be
be measured, place
nozzle in center or
or 3 feet into duct
Frequency and method
of measurement
Once each field test;
use wet bulb/dry bulb
thermometer, Method
4, or sling psychro-
meter
See above for
moisture content;
for flow rate, once
each field test using
Method 1 location,
if possible, and
Method 2 procedures
Prior to sampling,
using tape measure
Prior to sampling
Prior to sampling
Prior to sampling,
inspect all
connections
Prior to sampling,
determine using
stack dimensions
Action if require-
ments are not met
Complete
Complete
Complete
Complete
O
Repair or replace
if leaks found
Check for leaks,
repair system;
repeat check
Reposition
O
-------
Table 4.1 (Continued)
Section No. 3-17.4
Date May 31, 1991
Page 11
Characteristic
Acceptance limits
Frequency and method
of measurement
Action of require-
ments are not met
Sampling system
preparation
(continued)
Assemble system as
shown in Figure
1.1; no leakage
Check dry ice
level, and add
ice, if necessary
Close the purge
valve and stop
the pump; record
start time, open
sample valve and
sample tank, set
flow to calculated
flow rate and
record the tank
volume, flow rate
setting, probe
temp and filter
temp
Adjust flow rate
to maintain a
constant rate and
adjust probe and
temp controllers
as necessary to
maintain proper
temperatures
Record sample
tank vacuum, flow
meter settings,
probe temp and
filter temp at
5-min intervals
throughout sampling
Calculate sampling
rate for the test
Before sample collect-
ion visually and
physically inspect all
connections
Prior to sampling
To begin sampling
Check for leaks;
repair system;
repeat check
Complete
Complete
During sampling;
flow rate should
be kept at +_ 10%
throughout the
sampling period
During sampling
Repeat sampling
Complete
(Continued)
Prior to sampling;
divide sampling
time (min) into 80%
of the sample tank
volume (cc)
Complete
/ XI ,f
'' V<" '
-------
Section No. 3-17.4
Date May 31, 1991
Page 12
Table
(Continued)
o
Characteristic
Constant rate
sampling
(Continued)
Acceptance limits
Adjust probe tip
such that probe is
perpendicular to
stack axis- or nozzle
is located at a
point of average
velocity with tip
facing opposite of
direction of gas
flow
Set probe temp to
129°C; set filter
temp controller to
121°C allow probe
and filter to heat
for 30 min
Purge train, close
sample valve, open
purge valve, and
start vacuum pump
To begin sampling,
close purge valve
and stop pump; open
sample valve and
sample tank valve
Sampling must be
conducted at a
constant rate of
i 10% over duration
at a rate between
60 and 100 cc/min
Sample collection
ends when total
sampling time is
complete or constant
flow rate can no
longer be maintained
due to reduced
sample tank volume
Frequency and method
of measurement
Prior to sampling
Prior to purging the
train; confirm using
thermocouples
Before sample
collection, with
sample valve closed
During field test
During field test
During field test
Action of require-
ments are not met
Reposition; check
system for leaks
Adjust heating
system
Repeat purge
Repeat sampling
O
Repeat sampling
If sampling must be
interrupted before
reaching the
minimum sampling
time, see
Subsection 4.3.3,
Step 9
O
-------
Table 4.1 (Continued)
Section No. 3-17.4
Date May 31, 1991
Page 13
Characteristic
Acceptance limits
Frequency and method
of measurement
Action of require-
ments are not met
Constant rate
sampling
(continued)
After sampling is
completed, close
the flow control
valve; record final
tank vacuum and
temperature, and
barometric pressure;
close sample tank
valve
Conduct cylinder gas
performance audit
during sample
collection phase of
test
Immediately following
sampling
Complete
Collect two audit
sample gas concentra-
tions per test; see
Section 3.1?.8
Complete
Sample recovery
(Continued)
Disconnect sample
sample tank from
sampling system;
record tank pressure
within 1 mm Hg
Disconnect
condensate trap;
seal both ends
tightly
Identify condensate
trap and sample
tanks by their test
run number and
sampling location
and enter corres-
ponding information
on field data form
Pack trap samples
in dry ice during
storage and
shipping
Label clean sample
tank and condensate
as sample blanks
Following sampling
Complete
Following sampling
Following sampling
Complete
Complete
Following sample
recovery
Following sampling
Complete
Complete
-------
Section No. 3-17-
Date May 31, 1991
Page 14
o
Table 4.1 (Continued)
Characteristic
Sample logistics
Acceptance limits
Properly label all
bags , containers ,
etc; record all
data on form such
as Fig. 4.2
Frequency and method
of measurement
Visually check each
sample
Action of require-
ments are not met
Complete the
labeling
O
o
-------
Section No. 3.17-5
Date May 31, 1991
Page 1
5.0 POSTSAMPLING OPERATIONS
The postsampling operations include recovery of the condensible organic
fraction contained in the condensate trap, analysis of the carbon dioxide collected
during condensible organic recovery, and analysis of the noncondensible organic
fraction collected during source sampling in the sample tank. These operations are
performed for the compliance test samples, the blank sample trap/blank sample tank
set, and the audit samples.
Both initial and daily performance checks of the condensible organic recovery
system are performed. Initial performance checks include a carrier gas and
auxiliary oxygen blank test, an oxidation catalyst efficiency check, and a
condensible organic recovery efficiency test. Daily performance checks include a
recovery system leak check, a system background test, and an oxidation catalyst
efficiency check.
Both initial and daily performance checks and calibration of the NMO analyzer
are performed. Initial performance tests include an oxidation catalyst efficiency
check, a reduction catalyst efficiency check, an analyzer response linearity check,
and a chromatography performance check. The initial calibration is derived from
the analyzer linearity check data. Daily calibration of both the C02 response and
NMO response is performed. Figure 5-18 at the end of this section provides a
checklist for monitoring the postsampling operations. Table 5.1 at the end of the
section summarizes the quality assurance activities associated with the
postsampling operations.
5.1 Initial Performance Tests
Initial performance tests of both the condensible organic recovery system and
the NMO analyzer are performed before the systems are first placed into operation,
after any shutdown of longer than six months, or after any major modification of
the systems. ...
5.1.1 NMO Analyzer (see Figure 5.1)
Oxidation Catalyst Efficiency Test - With both catalysts unheated, perform
triplicate analyses of the high level methane standard (nominal 1 percent CHA in
air). With the oxidation catalyst heated only to its operating temperature,
reanalyze the high level methane standard in triplicate. Record data on a data
sheet (Figure 5-2) and calculate the oxidation catalyst efficiency using the
equation shown the figure. The average response with the oxidation catalyst heated
should be less than 1 percent of the average response obtained with both catalysts
unheated. If not, replace the oxidation catalyst.
Reduction Catalyst Efficiency Test - With the oxidation catalyst unheated and
the reduction catalyst heated to its operating temperature, analyze the high level
methane standard in triplicate. Repeat the analysis in triplicate with both
catalysts heated to their operating temperatures. Record data on a data sheet
(Figure 5-2) and calculate the reduction catalyst efficiency using the equation
shown the figure. The responses observed under these two conditions should agree
within 5 percent. If not, replace the reduction catalyst.
NMO Response Linearity Test and Initial Calibration - With both catalysts at
their operating temperatures, perform triplicate injections of each of the propane
-------
Section No. 3-17-5
Date May 31, 1991
Page 2
o
COLUMN OVEN
GC COLUMN
cgogggggoooooqcqcaxxjciD.
VALVE OVEN
Ho CARRIER
GAS
OXIDATION
CATALYST
Figure 5-l« Nonmethane organic analyzer.
O
-------
Section No. 3.17-5
Date May 31, 1991
Page 3
NMO Analyzer Catalyst Efficiency Testing
Date
Analyst
Oxidation Catalyst
Temp. , °C
\ '. v.
Reduction Catalyst
Temp., °C
FID Response
Run 1
Run 2
i .
Run 3
;• •
Average
t '
•<•- • , '• .
Oxidation Catalyst Efficiency =
Rl -
x 100 (criteria is 99# or greater)
where: Rj = Average FID response with both catalysts unheated.
R2 = Average FID response with oxidation catalyst only heated.
Reduction Catalyst Efficiency =
R,
x 100 (criteria is 95% or greater)
where: R3 = Average FID response with reduction catalyst only heated.
R4 = Average FID response with both catalysts heated.
Figure 5.2. Analytical data form for NMO analyzer catalyst efficiency.
-------
Section No. 3-17-5
Date May 31, 1991
Page 4
o
standards specified in Subsection 1.3.5 (i.e., 20 ppm, 200 ppm, and 3tOOO ppm in
air nominal). Convert certified concentrations in ppm to ppm C by multiplying ppm
concentrations by 3. Record these concentrations on a data sheet, such as shown
in Figure 5.3, along with the area responses observed in each injection. Calculate
the mean response factor as ppm C/mean area for each standard and the overall mean
response factor for all three standards. The NMO response linearity is acceptable
if the average response factor of each calibration gas standard is within 2.5
percent of the overall mean response factor and if the relative standard deviation
for each set of triplicate injections is less than 2 percent. If these criteria
are not met, check the air and hydrogen flows for the FID to confirm that they are
set according to manufacturer's specifications. Make adjustments if necessary and
repeat the test. The overall mean response factor is used as the initial NMO
calibration response factor (RFNMO).
C02 Response Linearity Test and Initial Calibration - Perform the linearity
test as described above, except use the C02 calibration standards specified in
Subsection 1.3-5 (50 ppm, 500 ppm, and 1 percent in air). The overall mean
response factor is used as the initial C02 calibration response factor (RFC02).
The C02 calibration response factor (RFC02) should be within 10 percent of the NMO
calibration factor (RFNMO). If not, repeat the oxidation catalyst efficiency test.
NMO Analyzer Performance Test - After calibration of the NMO response as
described above, analyze each of the test gas mixtures specified in Subsecti/ "\
1.3.5 in triplicate. (Standard 1 is nominally 50 ppm CO, 50 ppm CHA , 2 percertv_X
C02, and 20 ppm propane in air; Standard 2 is 50 ppm hexane in air; Standard 3 is
nominally 20 ppm toluene in air; and Standard 4 is 100 ppm methanol in air.)
Record the NMO area responses for each test gas on a data sheet such as shown in
Figure 5.4. Convert the certified organic compound concentrations of the test gas
mixtures to ppm C by multiplying by the carbon number of the compound (3 for
propane, 6 for hexane, and 7 f°r toluene). Record these concentrations on the data
sheet as the expected concentrations. Calculate the mean NMO concentration of the
test gas using the equation shown in Figure 5.4. The analyzer performance is
acceptable if the average measured NMO concentration for each mixture is within 5
percent of the expected value.
5.1.2 Condensible Organic "Recovery System
Carrier Gas and Auxiliary Oxygen Blank Check - Each new tank of zero grade
air and zero grade oxygen is analyzed with the NMO analyzer according to the
procedure described in Subsection 5.4.2. The total concentration from any
measured methane, carbon monoxide, carbon dioxide, or nonmethane organics should be
less than 5 parts-per-million carbon (ppm C). Record data on a data sheet such as
shown in Figure 5•5•
Oxidation Catalyst Efficiency Test - Perform this test using the equipment
shown in Figure 5-6 and the following procedure:
1. Install a clean sample trap in the recovery system.
2. Replace the zero air carrier gas with the high level methane S8?-*\
standard (1 percent methane in air nominal concentration). ( )
3. Set the 4-port valve to the trap recovery position and the sampTb /
recovery valve to the vent position. Establish a 100 cm3/minute flow of
the methane in air standard.
4. Attach an intermediate collection vessel (ICV) to the recovery system.
-------
Section No. 3.17-5
Date May 31, 1991
Page 5
NMO Analyzer Linearity Testing
Date
Analyst
Compound
Propane
Propane
Propane
C02
C02
C02
Cone . ,
ppm C
FID Area Response
Run 1
Run 2
Run 3
Mean
RSD
HP.
ppm C/Area
Overall
Mean RF
Percent
Difference
RF - Overall Mean RF
Percent Difference
x 100
Overall Mean RF
RSD =
100
••••—•••I
R
n - 1
where: R
Ri
n
Mean RID response.
FID response for run i,
Number of runs.
Mean RF =
Standard Cone, (ppm C)
Mean Area Response (R)
Overall Mean RF =
(RFNMO or RFC02)
RF2 + RF3
Figure 5.3. Analytical data form for NMO analyzer linearity tests.
-------
Section No. 3.17-5
Date May 31, 1991
Page 6
NMO Analyzer Performance Testing
o
Date
Analyst
Test Gas
Propane Mix
Hexane
Toluene
Methanol
NMO Area Response
Run 1
Run 2
--•*
Run 3
Mean
RFNMO •
ppm C/Area
Mean Cone . ,
ppm C
Expected
Cone. , ppm C
Percent
Diff.
Mean Cone. = Mean NMO Area x RF,
NMO
Percent Difference = Mean Cone. - Expected Cone.
Expected Cone.
O
Figure
Analytical data form for NMO analyzer performance test.
O
-------
Section No. 3-17-5
Date May 31, 1991
Page 7
Analysis of Recovery System Carrier Gases
Date
Analyst
Cylinder No.
Peak Area
CO C02 NMO
RF
fir,
NMO
Concentration, ppm C
CH4 CO C02 NMO
Concentration, ppm C = RFNMO x Peak Area
Total,
ppm C
Figure 5.5- Analytical data form for analysis of recovery system carrier gases.
-------
Section No. 3-17-5
Date May 31, 1991
Page 8
o
FLOW METERS
\
HEAT TRACE (100'C).
SAMPLE
RECOVERY
VALVE
FLOW
CONTROL
VALVE
SYRINGE PORT
O
VACUUM PUMP
Figure 5.6.
Condensible organic recovery system oxidation catalyst efficiency test.O
-------
7.
8.
Section No. 3.17-5
Date May 31, 1991
Page 9
With the flow control and ICV valves fully open, open the vacuum valve
to evacuate the manometer or gauge, connecting tubing, and the ICV to 10
mm Hg absolute pressure or less. Close the vacuum valve and flow
control valve.
Once the NDIR analyzer response is stable, switch the sample recovery
valve to the collect position. When the manometer or gauge begins to
indicate pressure above atmospheric, open the flow control valve to
maintain atmospheric pressure in the system. After fully opening the
flow control valve, continue pressurizing the ICV to a nominal gauge
pressure of 300 mm Hg (1060 mm Hg absolute).
Switch the sample recovery valve back to the vent position.
Close the ICV valve and detach the ICV from the system. Replace the
methane standard with the zero air carrier gas.
Analyze the C02 concentration in the ICV using the NMO analyzer. The
C02 concentration should be within 2 percent of the methane standard
concentration. Record data on a from such as that shown in Figure 5*7-
If the test criteria cannot be met, repack the oxidation catalyst tube
with new material as described in Subsection 1.2.1.
Condenstble Organic Recovery Efficiency Test - This test is performed
the equipment shown in Figure 5«8 and the following recommended procedure:
using
the
the
liquid sample injection unit in
end with the tee to the tubing
place of a sample
terminated at the
trap.
-port
to nominal
sample
5-
6.
8.
10.
11
12
is 100 i
each set
recovery
Install
Connect
valve.
Set the zero air carrier gas and auxiliary oxygen flows
levels of 100 cm3/minute and 150 cm3/minute, respectively.
Set the 4-port valve to the trap recovery position and the
recovery valve to the vent position.
Attach an intermediate collection vessel to the recovery system.
Evacuate the ICV, manometer or gauge, and connecting tubing to 10 mm Hg
absolute pressure or less. Close the vacuum and flow control valves.
Switch the sample recovery valve to the collect position.
Inject 50 microliters of hexane into the septum port of the liquid
sample injection unit.
Continue recovery of the injected organic as described in Subsection
5.3-2.
Record the final ICV volume and ICV pressure on a form such as that
shown in Figure 5-9 and detach the ICV from the recovery system.
Determine the C02 concentration of the ICV by analysis using the NMO
analyzer and record on a form such as that in Figure 5-9.
Calculate the percent recovery using the equation on a form such as that
shown in Figure 5.9.
Repeat the recovery test two additional times with 50 ul hexane
injections.
Perform additional recovery tests in triplicate with 10 ul hexane, 50 ul
decane, and 10 ul decane each.
The recovery system performance is acceptable if the average percent recovery
10 percent with a relative standard deviation of less than 5 percent for
of triplicate analyses. If these requirements are not met, check the
system for leaks and ensure adequate heating of the liquid sample
injection unit during recovery.
-------
Section No. 3.17.5
Date May 31, 1991
Page 10
Recovery System Oxidation Catalyst Testing
o
Date
Analyst
C02 Area Response
of ICV
i.
RFC02
ppm/area
C02 Cone . ,
ppm
Cfy Std.
Cone. , ppm
Percent
Difference
O
Percent Difference
°°?
Std> C°nC> X 10°
CHj| Std. Cone.
C02 Cone. = C02 Area Response x RFC02
Figure 5.7. Analytical data form for recovery system oxidation catalyst testing.
O
-------
Section No. 3.17-5
Date May 31, 1991
Page 11
FLOW METERS
\
HEAT TRACE (100°C)
AIR
LIQUID
SAMPLE
INJECTION
IMF—-OCtl
SAMPLE
RECOVERY
VALVE
FLOW
CONTROL
VALVE
SYRINGE PORT
CJ
VACUUM PUMP
ICV
CV
VALVE
Figure 5»8. Condensible organic recovery efficiency test.
-------
Condensible Organic Recovery Efficiency Testing
Date
Analyst
Compound
Hexane
Hexane
Hexane
Hexane
Hexane
Hexane
Decane
Decane
Decane
Decane
Decane
Decane
Molecular
Weight,
g/g-mole
86.18
86.18
86.18
86.18
86.18
86.18
142.29
142.29
142.29
142.29
142.29
142.29
Density,
g/ml
0.661
0.661
0.661
0.661
0.661
0.661
0.730
0.730
0.730
0.730
0.730
0.730
Volume
Injected,
uL
10
10
10
50
50
50
10
10
10
50
50
50
Collection Tank Data
Volume, Temp., Pressure,
m3 °K mm Hg
C02 Cone. ,
pprn
% Recovery
'•-
Overall Mean,
% Recovery
RSD
Percent
Percent Recovery = 1.604
(Molecular Weight)(Tank Volume)(Tank Pressure)(CO, Cone.)
(Volume Injected)(Liquid Density)(Tank Temp.)(Carbon Number of Compound Injected)
100
RSD =
n - 1
where: J!R1 = Percent recovery for an individual test.
%Ri = Overall mean percent recovery for a compound and injection volume.
n = Number of tests (3).
Figure 5.9. Analytical data form for recovery efficiency test.
O O
13 O Cfl
ffl P fi>
W rr O
(D 0> cr
^ 3 O
W» 3
"Z,
UO O
00
!-' •
V£> H»
VD-4
l-» •
VJl
O
-------
Section No. 3-17.5
Date May 31, 1991
Page 13
5.2 Daily Performance Tests
5.2.1 Condensible Organic "Recovery System {see Figure 5-10) - Complete three tests
each day before recovery of any samples.
Leak Test - Completely close the zero air and auxiliary oxygen metering valves
and the flow control valve. Install a clean sample trap in the system and switch
the sample recovery valve to the collect position. Open the vacuum valve and
evacuate the system to 10 mm Hg absolute pressure or less. Close the vacuum valve
and record the initial pressure on a data sheet such as shown in Figure 5.H-
After 10 minutes, record the system pressure. The pressure change should be less
than 2 mm Hg over the 10-minute period. If not, locate leakage by an appropriate
method such as pressurizing and checking fittings with water. Repair leaks and
retest as described above.
System Background Test - Set the carrier gas and auxiliary oxygen flows to
their normal values of 100 cm3/minute and 150 cm3/minute, respectively. Switch the
4-port valve to the trap recovery position and the sample recovery valve to the
vent position. Use a 10 cm3 syringe to withdraw a sample from the syringe port
located upstream of the NDIR analyzer. Flush the gas sampling loop on the NMO
analyzer with the syringe sample and then switch the sampling valve to inject.
Record the C02 peak area on a data sheet such as shown in Figure 5-12 and calculate
the C02 concentration. The system background is acceptable if the C02
concentration measured is less than 10 ppm. If the background concentration is
greater than 10 ppm, purge the recovery system with carrier gas and heat the trap
connecting tubing to remove residual organics, then repeat the background test.
Oxidation Catalyst Efficiency Test - Conduct this test as described in
Subsection 5-1•!• If the test criteria cannot be met, replace the oxidation
catalyst and retest.
5.2.2 NMO Analyzer Daily Calibration
C02 Response Calibration - Analyze the highest level (\%) C02 calibration
standard three times. Record the C02 peak areas on a data sheet such as that shown
in Figure 5.13 and calculate the average daily response factor (DRFC02). The
average daily response factor (DRFC02) should be within 5 percent of the initial
C02 calibration response factor (RFC02). If not, repeat the initial performance
test as described in Subsection 5.1.2 to establish a new RFC02. The daily response
factor is used to quantitate the C02 concentrations of the ICV samples.
NMO Response Calibration - Analyze the gas mixture containing nominal
concentrations of 50 ppm CO, 50 ppm CHA, 2 percent C02, and 20 ppm propane in air
in triplicate. Record the NMO peak areas on a data sheet such as that shown in
Figure 5'13 and calculate the average daily response factor (DRFNMO). The average
daily NMO response factor (DRFNMO) should be within 5 percent of the initial NMO
response factor (RFNMO). If not, repeat the initial performance test as described
in Subsection 5«1«2 to establish a new RFNMO. The daily response factor is used to
quantitate the NMO concentration of the sample tanks.
-------
Section No. 3.17.5
Date May 31, 1991
Page 14
o
FLOW METERS
HEAT TRACE (100°C).
SAMPLE
RECOVERY
VALVE
aow
CONTROL
VALVE
SYRINGE PORT
O
VACUl/M PUMP
Figure 5.10. Condensate recovery system.
O
-------
Section No. 3.17-5
Date May 31, 1991
Page 15
Condensate Recovery System Leak Testing
Date
Analyst
Initial
Time
Initial Press.,
Torr
Final Time
Final Press.,
Torr
Figure 5.11. Analytical data form for recovery system leak test.
-------
Section No. 3.17-5
Date May 31, 1991
Page 16
Condensate Recovery System Background Testing
Date
Analyst
RFC02 •
ppm/Area
C02 Peak Area
CO, Cone., ppm
o
o
C02 cone., ppm = CO- Peak Area x RF,
C02
Figure 5.12. Analytical data form for recovery system background test.
O
-------
Section No. 3.17.5
Date Hay 31, 1991
Page 17
Daily NMO Analyzer Calibration
Date
Analyst
NMO Calibration Cylinder No.
CO, Calibration Cylinder No.
Compound
Cone . ,
ppm C
FID Area Response
Run 1
Run 2
Run 3
Mean
DRF,
ppm C/Area
Initial
RF
Percent
Diff.
Percent Difference = DRF " Initial RFx 100
Initial RF
R
Mean Area Response =
where: RA = FID area response.
Concentration
DRF =
Mean Area Concentration
Figure 5-13. Analytical data form for daily calibration of NMO analyzer.
-------
Section No. 3-17.5
Date May 31. 1991
Page 18
5.3 Condensible Organic Fraction Recovery
o
Recovery of condensible organics is accomplished in two steps. First, the
condensate trap is purged of carbon dioxide while cooling the trap in dry ice.
Second, the condensible organics are volatilized and converted catalytically to
carbon dioxide which is collected in an intermediate collection vessel (ICV) for
analysis.
5.3.1 Trap Purge and Sample Tank Pressurizatton - The following procedure is used
to purge carbon dioxide from the condensate trap and to pressurize the sample tank
(see Figure 5.1*0:
1. Obtain the sample tank and condensate trap from the source test run to be
analyzed (or the blank' sample tank and blank condenste trap).
2. Set the zero air carrier gas to a flow rate of 100 cm3/minute and the
auxiliary oxygen flow to zero.
3. Switch the *J-port valve to the C02 purge position.
4. Attach the sample tank to the condensate trap recovery system.
5. With the sample recovery valve in the vent position and the flow control
valve fully open, evacuate the manometer or gauge to the expected pressure of
the sample tank.
6. Close the vacuum valve, open the sample tank valve, and record the sample
tank pressure (Pt) in mm Hg absolute on a data sheet such as that shown in
Figure 5.15.
7. Immerse the condensate trap in crushed dry ice and attach to the recovery^—^
system with the trap outlet connected to the tube terminating at the k-port( j
valve. V /
8. Switch the sample recovery valve from vent to collect. Adjust the flow
control valve to maintain approximately atmospheric pressure in the recovery
system.
9. After the NDIR analyzer responds to the C02 purged from the trap and the
response reaches a minimum level, withdraw a 10 cm3 syringe sample from the
syringe port and analyze with the NMO analyzer. Repeat analyses until the
C02 concentration of the trap effluent is less than 5 ppm. The length of
time required to purge the trap of residual C02 will depend upon the internal
volumes of the condensate trap recovery system. A larger volume system will
require more purging time at the specified flow rates to meet the effluent
concentration criteria of less than 5 ppm.
10. Switch the carrier gas bypass valve to pressurize the sample tank to
approximately 1060 mm Hg absolute pressure. Switch the sample recovery valve
to the vent position and record the final sample tank pressure (Ptf) on the
data sheet. Detach the sample tank from the system.
5-3'2 "Recovery of Condensible Organics - The following procedure is used to purge
organics from the sample trap, convert them to carbon dioxide, and collect the
carbon dioxide in an intermediate collection vessel (see Figure 5.16).
1. Attach an ICV to the trap recovery system. Open the flow control and ICV
values fully and evacuate the manometer or gauge, connecting tube, and ICV to
10 mm Hg absolute pressure or less. Close the flow control and vacuum
valves.
2. Set the auxiliary oxygen flow to a rate of 150 cm3/minute. /"""N
3. Switch the 4-port valve to the trap recovery position and the sample ( J
recovery valve to the collect position. After the system reaches ^—
atmospheric pressure, adjust the flow control valve to maintain atmospheric
pressure within 10 percent.
-------
Section No. 3.17.5
Date May 31, 1991
Page 19
FLOW METERS
\
HEAT TRACE (100°C)
SAMPLE
RECOVERY
VALVE
FLOW
CONTROL
VALVE
SYRINGE PORT
VACUUM PUMP
SAMPLE
TANK
Figure 5.14. Condensate recovery system, CO purge.
-------
Section No. 3.17-5
Date May 31, 1991
Page 20
Date
Location
Run No.
Analyst
Plant _
Date Sampled
o
Trap No.
Sample Tank No.
ICV No.
ICV volume, m3 (Vv)
Sample tank pressure after sampling, mm Hg (Pt)
Sample tank pressure after pressurizing, mm Hg {Ptf)
ICV final pressure, mm Hg (Pf)
ICV volume, m3 (Vv)
Sample tank temperature after pressurizing, °K (Ttf)
ICV final temperature, °K (Tf) ,
Sample tank temperature at end of sampling, °K (tt)
Sample tank temperature before sampling, °K (Tti)
Sample tank pressure before sampling, mm Hg (Ptl)
Gas volume sampled, dsm3 (Vg)
Run
No.
1
2
3
Mean
ICV Analysis
C02 Area
C02 Cone . ,
PPm (Ccnl)
NMO Area
NMO Cone . ,
ppm C (Ccm2)
,
Sample Tank Analysis
NMO Area
NMO Cone . ,
PPm c (ctm)
o
C02 Cone. = C02 Area x
NMO Cone. = NMO Area x DRFMMn
ri jn u
rtf
T
Atf
Noncondensible Organic = -= =—
Concentration, ppm C (Ct) _J_ t^
IT—x ct»
V P
vv rf
x <% + Ccm2 )
Condensible Organic = 0.3857
Concentration, ppm C (Cc)
TGNMO Concentration, ppm C = Ct - Ctb + Cp - Ccb*
*Note: Blank subtraction must have prior approval of the Administrator.
Figure 5-15- Analytical data form for sample recovery and analysis.
O
-------
Section No. 3.17-5
Date May 31, 1991
Page 21
FLOW METERS
HEAT TRACE (100°C).
SAMPLE
RECOVERY
VALVE
FLOW
CONTROL
VALVE
SYRINGE PORT
VACUUM PUMP
Figure 5*16. Condensate recovery system, collection of trap organics,
-------
o
Section No. 3-17.5
Date May 31, 1991
Page 22
4. Remove the condensate trap from the dry ice and allow ambient air warming of
the trap while monitoring the NDIR analyzer response. The auxiliary oxygen
flow may be discontinued after five minutes if the C02 response is below
10,000 ppm (1%).
5. Heat the trap by placing it in a furnace at 200° C. If the NDIR response
exceeds 50,000 ppm (5#) during recovery, resume auxiliary oxygen flow at a
rate of 150 cm3/minute.
6. After the NDIR analyzer indicates a C02 concentration of less than 10,000
ppm, begin heating the tubing connecting the condensate trap to the
oxidation catalyst with a heat gun. Heat the tubing slowly along the entire
length from the trap 'to tfhe catalyst, and repeat two additional times.
7. Continue trap heating and purging until the C02 concentration is below 10
ppm (determined by analyzing syringe samples collected before NDIR analyzer
with the NMO analyzer).
8. When recovery is complete, switch the carrier gas bypass to pressurize the
ICV to approximately 1060 mm Hg. Switch the sample recovery valve to vent
and record the ICV final pressure (Pf) on a data sheet such as that shown in
Figure 5-15.
5-4 Analysis
The total source concentration of gaseous nonmethane organics is determined by
combining the noncondensible and condensible concentrations. The noncondensible
concentration is determined by analyzing the sample tank for nonmethane organics,
and the condensible concentration is determined by analyzing the intermediate
collection vessel for C02.
5.4.1 NMO Analyser Operating Conditions - Set the helium carrier gas flow rate to
30 cm3/minute. Set the oxidation catalyst oxygen flow rate to 2.2 cm3/minute.
Heat the column oven to an initial temperature of 85°C.
5.^.2 Intermediate Collection Vessel Analysts - Analyze the ICV contents as
follows:
1. Attach the ICV to the 10-port gas sampling valve.
2. Purge the sample loop with gas from the ICV and then switch the 10-port valve
to the inject position.
3. When the detector response returns to near baseline following the C02 peak,
switch the 10-port valve to the backflush position and increase the column
oven temperature to 195°C as rapidly as possible.
4. After detection of any nonmethane organic compounds, return the column oven
temperature to 85°C.
5. Record the C02 peak area and NMO peak area on the data sheet shown in Figure
5.15-
6. Repeat the analysis two additional times.
7. Calculate C02 and NMO concentrations using equations given in Figure 5.15.
8. Calculate the average C02 concentration (Cc(nl) and the average NMO
concentration (Ccin2) in the ICV and record on the data sheet.
5.4.3 Sample Tank Analysis - Analyze the sample tank as described in Section
5.4.2. Record the NMO peak area only and calculate the average NMO concentration f-^.
(Ctra) of the sample tank. f j
o
-------
Section No. 3-17-5
Date May 31. 1991
Page 23
5.4.4 Condensible Organic Blank Analysis -
1. Analyze the ICV resulting from the blank trap recovery as described in
Section 5.4.2.
2. Calculate C02 and NMO concentrations using equations given in Figure
5.16.
3. Calculate the average C02 concentration (Ccnl) and the average NMO
concentration (Ccm2) in the blank ICV and record on the data sheet.
4. Calculate the source condensible organic blank concentration (Ccb) using
the equation in Figure 5-17- The condensible organic blank concentration
may not exceed 15 ppmC. If the blank value exceeds 15 ppmC, then the
value of 15 ppmC may be used as the blank value. NOTE: The method does
not provide for blank correction. However, with prior approval of the
Administrator, blank correction (subtracting the blank value) may be
used.
5-4.5 Noncondensible Organic Blank Analysis - Analyze the blank sample tank as
described in Section 5-4.2. Record the NMO peak area only and calculate the
average NMO concentration (Ctmb)of the blank sample tank. Calculate the source
noncondensible organic blank concentration (Ctb) using the equation shown in
Figure 5-17- The noncondensible organic blank concentration may not exceed 5
ppmC. If the blank value exceeds 5 ppmC, then the value of 5 PpmC may be used as
the blank value. NOTE: The method does not provide for blank correction.
However, with prior approval of the Administrator, blank correction (subtracting
the blank value) may be used.
5.4.6 Source Concentration Calculations - Calculate the noncondensible organic
concentration (Ct), the condensible organic concentration (Cc), and the TGNMO
concentration using the data and equations shown Figures 5.15 and 5.17 (blank
subtraction requires prior approval of Adminstrator).
-------
Section No. 3-17-5
Date Hay 31, 1991
Page 24
Date
Location
Blank No.
Sample Tank No.
Analyst
Plant _
Date Sampled
o
Trap No.
ICV No.
ICV volume, o3 (Vv)
Sample tank pressure after sampling, nun Hg (Pt)
Sample tank pressure after pressurizing, mm Hg (Pt t)
ICV final pressure, mm Hg (Pf)
ICV volume, ra3 (Vv)
Sample tank temperature after'ipressurizing, °K (Ttf)
ICV final temperature, °K (Tfj
Sample tank temperature at end of sampling, °K (Tt)
Sample tank temperature before sampling, °K (Ttl)
Sample tank pressure before sampling, mm Hg (Ptl)
Gas volume sampled, dsm3 (VB)
Blank
No.
1
2
3
Mean
ICV Analysis
C02 Area
C02 Cone . ,
PPm (Ccml)
NMO Area
NMO Cone . ,
PPm c
-------
Section No. 3.17-5
Date May 31, 1991
Page 25
Date Plant Name Sampling Location
Initial Performance Tests of Condensible Organic Recovery System
Do zero grade air and zero grade oxygen carrier gases contain less than 5 Ppm C
total of methane, carbon monoxide, carbon dioxide, and nonmethane organics?
yes no. (If no, replace gases or filter to remove impurities.)
Does the C02 concentration in the ICV collected during the oxidation catalyst
efficiency test agree within 2 percent of the methane standard concentration?
yes no. (If no, replace the oxidation catalyst.)
Is recovery efficiency 100 +_ 10 percent with a relative standard deviation of less
than 5 percent for each set of triplicate injections of hexane and decane at
10 ul and 50 ul? yes no. (If no, check the recovery system for
leaks and assure adequate heating of the liquid sample injection unit.)
Initial Performance Tests of NMO Analyzer
Does the oxidation catalyst efficiency test show an efficiency of 99 percent or
better? yes no. (If no, replace the oxidation catalyst.)
Does the reduction catalyst efficiency test show an efficiency of 95 percent or
better? yes no. (If no, replace the reduction catalyst.)
Is the NMO response linear? (Average response factor of each calibration standard
within 2.5 percent of the overall mean and a relative standard deviation for
each set of triplicate injections of less than 2 percent.) yes no.
(If no, check FID air and hydrogen flows and make adjustments.)
Is the C02 response linear? (Average response factor of each calibration standard
within 2.5 percent of the overall mean and a relative standard deviation for
each set of triplicate injections of less than 2 percent). yes no.
(If no, check FID air and hydrogen flows and make adjustments.)
Are the measured concentrations of the analyzer performance test gases within 5
percent of their certified concentrations? yes no. (If no, replace
the GC column or adjust column backflush timing and/or temperature to obtain
acceptable performance.)
Daily Performance Tests of Condensible Organic Recovery System
Leak test completed? yes no. (Leak test following procedure in Sub-
section 5.2.1.)
System background test completed? yes no. (Follow procedure in Sub-
section 5.2.1.)
Oxidation catalyst efficiency test completed? yes no. (Follow procedure
in Subsection 5.2.1.)
Daily Calibration of NMO Analyzer
Calibration of C02 response completed with 1% C02 in air standard?
yes no.
Calibration of NMO response completed with mixed gas standard containing 50 ppm
CO, 50 ppm CH4, 2 percent C02, and 20 ppm propane in air? yes no.
Figure 5.18. Postsampling operations checklist.
(Continued)
-------
Section No. 3.17-5
Date Hay 31, 1991
Page 26
Figure 5.18. (Continued)
Condensible Organic Recovery
Condensate trap purged and sample tank pressurized? yes no. (Follow
procedure in Subsection 5 • 3 • 1 •)
Condensible organics purged from sample trap, converted to C02, and collected in
an ICV? yes no. (Follow procedure in Subsection 5-3-2.)
Analysis of Sample Tank and Intermediate Collection Vessel
j . .,^ . ;
NMO analyzer operating conditions set? yes no. See Subsection,5-4.1..)
ICV analyzed in triplicate? yes no. (Follow procedure in Subsection
5.4.2.)
Sample tank analyzed in triplicate? yes no. (Follow procedure in
Subsection 5.4.3.)
Source concentration calculated? yes no. (Use equations shown in
Figure 5.1.5.)
o
o
-------
Section No. 3.17.5
Date May 31, 1991
Page 27
Table 5.1. ACTIVITY MATRIX FOR SAMPLE ANALYSIS
Characteristic
Carrier gas and
auxiliary 02 used
in condensible
organic recovery
system
Recovery system
oxidation catalyst
efficiency
Acceptance
limits
Total concentration
from CH4, CO, C02,
and NMO must be
<5 ppm C
C02 concentration
of collected sample
must be +. 2% of CHA
test gas concentra-
tion
Frequency and method
of measurement
Analyze each new
cylinder with the NMO
analyzer
Before first operation
after any shutdown of
longer than 6 months,
after modification, or
daily when samples are
analyzed; test by re-
placing carrier gas
with 1% din in air,
collection of sample
in ICV, and analysis
with NMO analyzer
Action if
requirements
are not met
Obtain better
grades of gas
from vendor or
filter gases to
reduce impurities
Replace oxida-
tion catalyst
Condensible organ-
ic recovery
efficiency
Average recovery of
100 + 102 with a RSD
of <5# for each set
of triplicate analy-
ses
Before first operation
after any shutdown of
longer than 6 months,
or after modification;
test with 10 and 50 ul
injections of hexane
and decane into liquid
sample injection unit
and analysis of col-
lected sample with
NMO analyzer
Check recovery
system for leaks
and assure
adequate heating
of liquid sample
injection unit
during recovery
NMO analyzer oxi-
dation catalyst
efficiency
FID response with
oxidation catalyst
heated must be <1%
of response with
both catalysts un-
heated
Before first operation
after any shutdown of
longer than 6 months,
or after modification;
analyze 1% CHA in air
with both catalysts
unheated and then with
oxidation catalyst
only heated
Replace
oxidation
catalyst
(Continued)
-------
Table 5.1 (Continued)
Section No. 3.17.5
Date May 31. 1991
Page 28
o
Characteristic
Acceptance
Limits
Frequency and method
of measurement
Action if
requirements
are not met
NMO analyzer
reduction
catalyst
efficiency
Analyzer response
to 1% methane in air
standard with both
catalysts heated
should agree +_ 5% of
response, with re-
duction catalyst
only heated
Before first operation
after any shutdown of
more than 6 months, or
after modification; a
1% CH4 in air standard
is analyzed with both
catalysts heated and
with the reduction
catalyst only heated
Replace
reduction
catalyst
NMO response lin-
earity and cali-
bration
Response factor of
each calibration
gas standard must be
+2.5% of overall
mean response factor
and relative stan-
dard deviation of
each set of trip-
licate analyses
must be <2%
Before first operation
after any shutdown of
longer than 6 months,
or after modification;
analyze propane in air
standards with NMO
analyzer under normal
operating conditions
Check air and
hydrogen flows
for FID to con-
firm that they
are set accord-
ing to manufac-
turer's speci-
fications ; make
adjustments if
necessary and
and repeat test
O
C02 response
linearity and
calibration
Response, factor of
each calibration
gas standard must
be +2.5% of overall
mean response factor
and relative stan-
dard deviation of
each set of tripli-
cate analyses must
be < 2%
Before first operation
after any shutdown of
longer than 6 months,
or after modification;
analyze the C02 stan-
dards with NMO analy-
zer under normal oper-
ating conditions
Check air and
hydrogen flows
for FID to con-
firm they are
set according
to manufacturer's
specifications;
make adjustments
if necessary and
repeat test
NMO analyzer
performance
Average concentra-
tion based on trip-
licate analysis
must be within 5%
of expected value
for each test
mixture
Before first operation
after any shutdown of
longer than 6 months,
or after modification;
analyze test mixtures
with NMO anlayzer
under normal operating
conditions
Replace GC
column or
adjust column
backflush timing
and/or tempera-
ture to obtain
acceptable
performance
(continued)
O
-------
TABLE 5.1 (continued)
Section No. 3-17-5
Date May 31, 1991
Page 29
Characteristic
Condensible
organic recovery
system leak test
Recovery system
background test
Acceptance
Limits
System should lose
<2 mm Hg vacuum
over a 10-min period
The measured C02
background concen-
tration must be <10
ppm
Frequency and method
of measurement
Daily, before analysis
of samples; evacuate
the recovery system
and monitor the vacuum
with mercury manometer
or pressure gauge
Daily, before analysis
of samples; analyze
syringe samples of
.recovery system efflu-
ent with NMO analyzer
Action if
requirements
are not met
Locate leakage
by appropriate
method such as
pressurizing
and checking
fittings with
water; repair
leaks and retest
Purge recovery
system with
carrier gas and
heat trap con-
necting tubing
to remove res-
idual organics,
then retest
C02 response
calibration
The average response
factor from tripli-
cate analysis of
highest concentra-
tion standard must
be +$% of initial
C02 response factor
Before and after anal-
ysis of each set of
samples or daily,
whichever occurs
first, analyze high-
est level C02 standard
with NMO analyzer
under normal operating
conditions
Repeat analyzer
catalyst effic-
iency tests,
linearity tests,
and performance
test
NMO response
calibration
Average NMO response
factor should be +
of initial NMO
response factor
Before and after anal
ysis of each set of
samples or daily,
whichever occurs
first, analyze gas
mixture containing
50 ppm CO, 50 ppm CH4,
2% C02, and 20 ppm
propane in air
Repeat analyzer
catalyst effic-
iency tests,
linearity tests,
and performance
test
Sample trap
purge and
sample tank
pressurization
(Continued)
C02 concentration of
syringe port samples
must be less than
5
Analyze syringe port
samples with NMO
analyzer after NDIR
analyzer response
returns to baseline
Continue purg-
ing trap with
carrier gas and
analyze addi-
tional syringe
port samples
-------
Section No. 3.17.5
Date May 31. 1991
Page 30
TABLE 5.1 (continued)
o
Characteris tic
Acceptance
Limits
Frequency and method
of measurement
Action if
requirements
are not met
Condensible
organics recovery
C02 concentration of
syringe port samples
must be <10 ppm
Analyze syringe port
samples with NMO anal-
yzer when NDIR anal-
yzer response returns
to baseline
Continue heating
of sample trap
and purging with
carrier gas
Intermediate
collection vessel
analysis
Relative standard
deviation from trip-
licate analysis for
C0 and NMO must be
Analyze ICV after trap
recovery by injecting
aliquots on NMO
analyzer
Perform addi-
tional analyses
until RSD of
last three
injections
is <2%
Sample tank
analysis
Relative standard
deviation from
triplicate analysis
for NMO must be <2%
Analyze sample tank
after trap purging and
tank pressurization by
injecting aliquots on
NMO analyzer
Perform addi-
tional analyses
until RSD of
last three
injections is
O
O
-------
Section No. 3.17.6
Date May 31, 1991
Page 1
6.0 CALCULATIONS
Calculation errors due to procedural or mathematical mistakes can be a part of
total system error. Therefore, it is recommended that each set of calculations be
repeated or spotchecked, preferably by a team member other than the one who per-
formed the original calculations. If a difference greater than typical round-off
error is detected, the calculations should be checked step-by-step until the source
of error is found and corrected.
Calculations should be carried out to at least one extra decimal figure beyond
that of the acquired data and should be rounded off after final calculation to two
significant digits for each run or sample. All rounding of numbers should be
performed in accordance with the ASTM 380-76 procedures. All calculations should
then be recorded on a calculation form such as Figure 6.1.
A computer program is advantageous in reducing calculation errors. If a
computer program is used, the original data entered should be included in the
printout so it can be reviewed; if differences are observed, a new computer run
should be made. A computer program is also useful in maintaining a standardized
format for reporting of results. It is highly recommended that a standardized
format including the data shown in Figure 6.2 be used for reporting the emissions
results. The data shown will allow auditing of the calculations.
Table 6.1 at the end of this section summarizes the quality assurance
activities for calculations.
6.1 Nomenclature
The following nomenclature is used in the calculations:
C = TGNMO concentration of the effluent, ppm C equivalent.
Cc = Calculated condensible organic (condensate trap) concentration of the
effluent, ppm C equivalent.
Ccl) = Calculated condensible organic (condensate trap) blank concentration
of the sampling equipment, ppm C equivalent.
C = Measured concentration (NMO analyzer) for the condensate trap ICV, ppm
CO
'2-
C
C
C
C
L. m
NMO.
F = Sampling flow rate, cc/min.
L = Volume of liquid injected, ul.
M = Molecular weight of the liquid injected, g/g-mole.
mc = TGNMO mass concentration of the effluent, mg C/dsm3.
cob= Measured blank concentration (NMO analyzer) for the condensate trap
ICV, ppm C02.
t = Calculated noncondensible organic concentration (sample tank) of the
effluent, ppm C equivalent.
tb = Calculated noncondensible organic blank concentration (sample tank) of
the sampling equipment, ppm C equivalent.
tm = Measured concentration (NMO analyzer) for the sample tank, ppm
NMO.
tm = Measured blank concentration (NMO analyzer) for the sample tank, ppm
-------
Section No. 3.17-6
Date May 31, 1991
Page 2
o
N = Carbon number of the liquid compound injected (N = 12 for decane, N =
6 for hexane).
Pf = Final pressure of the intermediate collection vessel, mm Hg absolute,
Pb = Barometric pressure, cm Hg.
Ptl = Gas sample tank pressure before sampling, mm Hg absolute.
Pt = Gas sample tank pressure after sampling, but before pressurizing, mm
Hg absolute.
Ptf = Final gas sample tank pressure after pressurizing, mm Hg absolute.
Tf = Final temperature of intermediate collection vessel, °K.
Ttl = Sample tank temperature before sampling, °K.
Tt = Sample tank temperature at completion of sampling, °K.
Ttf = Sample tank temperature after pressurizing, °K.
V = Sample tank volume, m3.
Vt = Sample train volume, cc.
Vv = Intermediate collection vessel volume, m3.
VB = Gas volume sampled, dsm3.
n = Number of data points.
q = Total number of analyzer injections of intermediate collection
vessel during analysis (where k = injection number, 1 ... q).
r = Total number of analyzer injections of sample tank during analysis
(where j = injection number, 1 ... r).
XL = Individual measurements.
x = Mean value. x""X
P = Density of liquid injected, g/cc. j }
0 = Leak check period, min. V, /
AP = Allowable pressure change, cm Hg.
6.2 Calculations
The following are the equations used with the example calculation form, Figure
6.1 to calculate the concentration of TGNMO, the allowable limit for the pretest
leak check, and assess the efficiency of the condensate recovery system.
6.2.1 Allowable Pressure Change - Calculate the allowable pressure change, in cm
Hg, for the pretest leak check using the following equation. This value is then
compared to the actual pressure change, in cm Hg, to determine if the train is
suitable for sampling.
Equation 6-1
6.2.2 Sample Volume - For each test run, calculate the gas volume sampled using
the following equation.
Vs = 0.3857 V
p P
rt _ rti
T7 T77
Equation 6-2
6.2.3 Noncondenslble Organtcs Concentration - For each sample tank, determine the
concentration of nonmethane organics, in ppm C, using Equation 6-3.
-------
Section No. 3.17-6
Date May 31, 1991
Page 3
6.2.3 ffoncondensible Organics Concentration - For each sample tank, determine the
concentration of nonmethane organics, in ppm C, using Equation 6-3.
ptr
Ttf
Pt Ptl
Tt Ttl
1 V Cfc
— z_j tm
I* . *
Equation 6-3
6.2.4 Noncondensible Organics Blank Concentration - For blank sample tank,
determine the concentration of nonmethane organics, in ppm C, using Equation 6-3
and the values for Ctnb. The blank value may not exceed 5 PPm. If the blank
value exceeds 5 ppm C, then the value of 5 PPm C may be used as the blank value.
The calculated blank value is C
t b
6.2.5 Condensible Organics Concentration - For each condensate trap, determine the
concentration of organics, in ppm C, using Equation 6-4.
cp = 0.3857
q
i Z
q k=l
cm.
Equation 6-4
6.2.6 Condensible Organics Concentration - For each condensate trap, determine the
concentration of organics, in ppm C, using Equation 6-4 and the values for CCBb.
The blank value may not exceed 15 ppm. If the blank value exceeds 15 ppm C, then
the value of 15 ppm C may be used as the blank value. The calculated blank value
is Ccb.
6.2.7 TGNMO Concentration - To determine the TGNMO concentration for each test
run, use Equation 6-5. NOTE: The method does not provide for blank correction.
The tester must have prior approval of the Administrator to use blank correction
(subtract blank).
= ct - ctb + cc - ccb
Equation 6-5
6.2.8 TGNMO Mass Concentration - To determine the TGNMO mass concentration as
carbon for each test run, use Equation 6-6.
m, = 0.4993 C
Equation 6-6
• }
-------
Section No. 3.17.6
Date May 31. 1991
Page ^
6.2.9 Percent Eecovery - Calculate the percent recovery for the liquid organic
injections used to assess the efficiency of the condensate recovery and
conditioning system using Equation 6-7. The average recovery for triplicate
injections shoul fall within 102 (of 100/0.
Percent Recovery = 1.604 —
L P Tf N Equation 6-7
\ v •
6.2.10 "Relative Standard Deviation - Calculate the relative standard deviation
(RSD) for the percent recoveries for triplicate injections of liquid organics using
Equation 6-8. The RSD should be less than 5% for each set of triplicate analyses.
o
100 /^(Xi - x)2
RSD = ^ I Equation 6-8
~ V n - 1
x »
O
o
-------
Section No. 3.17.6
Date May 31, 1991
Page 5
F
V.
ALLOWABLE PRETEST LEAK CHECK PRESSURE CHANGE
cc/min, Pb = . cm Hg, 0 = . min,
cc
AP = 0.01
= . cm Hg
Equation 6-1
Si i-
I'
1.
V = 0. m3, Ptl
Pti = . mm Hg, Tj =
K vs = 0.3857 V
/
Pt _ pti
Tt Ttl
SAMPLE VOLUME
= . mm Hg,
. °K, Tt
= 0.
dsm3
>K
Equation 6-2
pti "
Ttl
Ctml
NONCONDENSIBLE ORGANICS CONCENTRATION
mm Hg, Pt = mm Hg, Ptf = mm Hg,
°K T - °K T - <
• __ *»• A£ __ __ > *^ > •'•tf °~ — *
. ppm NMO, Ctm2 = . ppm NMO,
. ppm NMO, r =
K,
tf
P P
rt rt
T T
x x
*.
tm
- . ppm C
Equation 6-3
Figure 6.1. Calculation form for Method 25 analysis.
-------
Section No. 3-17.6
Date May 31, 1991
Page 6
o
P'i =
Lti
'tnbl
NONCONDESIBLE ORGANICS BLANK CONCENTRATION
__ mm Hg, Pt = ___ mm Hg, Ptf =
__ . _ °K, Tt = ___ . _ °K , Ttf =
___ . _ ppm NMO, Ctmb2 = ___ . _ ppm NMO,
___ . _ ppm NMO, r = _
mm Hg,
'tb
t t
= . ppm C
Equation 6-3
CONDENSIBLE ORGANICS CONCENTRATION
Vv = 0. m3, V. = 0. dsm3,
Pf = mm Hg, Tf = . °K,
= . ppm C02 , Ccm2 = . ppm C02 ,
O
cm3
= . ppm C02 , q =
Cc = 0.3857
cm,.
Equation 6-4
= . ppm C
Figure 6.1. Continued
O
v --1 - -
-------
Section No. 3-17.6
Date May 31, 1991
Page 7
Vv = 0.
'crabl
CONDENSIBLE ORGANICS BLANK CONCENTRATION
m3, V8 = 0. dsm3,
mm Hg, Tf = . _.°K,
. _ ppm C02 , Ccmb2 = . _ ppm C02
. ppm C02, q =
V.
Crh = 0.3857
'cb
cm
k=l
bk
Equation 6-4
= . ppm C02
TGNMO CONCENTRATION
Ct = . ppm C, Ctb = . ppm C, Cc = . ppm C,
Ccb = . ppm C
c • ct ~ ctb*
cc " ccb = ._ ppm C
C =
Equation 6-5
Note: Blank subtraction requires prior approval of the Adminstrator.
TGNMO MASS CONCENTRATION
. ppm C
mc = 0.4993
. mg C/dsm3
Equation 6-6
Figure 6.1. Continued
-------
Section No. 3.17.6
Date May 31, 1991
Page 8
PERCENT RECOVERY
. g/g-mole, L = . ul,
. mm Hg, Tf = . °K,
Vv = 0. . m3 , p = 0. kg/cc,
Ccm.= . ppm C, N =
o
pf ccm
^ I V J. w III
Percent Recovery = 1.604 — =
Tf N
Equation 6-7
RELATIVE STANDARD DEVIATION
n =
100
RSD =
n - 1
Equation 6-8
O
Figure 6.1. Continued
O
-------
FIELD DATA AND RESULTS TABULATION
Plant: XXXXXXXXXX
Sampling Location: XXXXXXXXXX
Date
Run Start Time
Run Finish Time
Field Data
Sample Trap I.D.
Sample Tank I.D.
Sample Tank Volume, V (n^)
Actual Volume Sampled, V, (dsm^*)
s
Field Initial Barometric Pressure, Pb (cm Hg)
Field Final Barometric Pressure (cm Hg)
Run 1 Run 2 Run 3 Blank Audit 1 Audit 2
xx/xx/xx xx/xx/xx xx/xx/xx xx/xx/xx xx/xx/xx xx/xx/xx
Field Initial Gauge Pressure of Tank,
Field Final Gauge Pressure of Tank
nk, Ptl
. pt U
(mm Hg absolute)
mm Hg absolute)
Field Initial Temperature of Tank, Tt; (
Field Final Temperature of Tank, Tt (*K)
Laboratory Data
Final Tank Pressure, Pfcf (mm Hg absolute)
Final Tank Temperature, Ttf (°K)
Noncondenslble (tank) Portion - Injection #1 (area units)
Noncondensible (tank) Portion - Injection #2 (area units)
Noncondensible (tank) Portion - Injection #3 (area units)
Instrument Blank (area units)
NMO Response Factor (area units/ppm C)
•68°F -- 29.92 in. Hg (760 ran Hg)
T) O C/5
(0 {D 0>
Oq rt O
(D (D ft
(continued)
Figure 6.2. Recommended standard format for reporting Method 25 data and results.
U> O
I-* •
LO
I-' •
MD—J
ON
-------
Run 1
Run 2
Run 3
Blank
Audit 1
Audit 2
xx/xx/xx xx/xx/xx xx/xx/xx xx/xx/xx xx/xx/xx xx/xx/xx
Laboratory Data (Continued)
Volume of ICV, Vy (n3)
Final ICV Pressure, P- (ran Hg absolute)
Final ICV Temperature. Tf (°K)
Condensible (trap) Portion - Injection fi (area units)
Condensible (trap) Portion - Injection #2 (area units)
Condensible (trap) Portion - Injection #3 (area units)
Instrument Blank (area units)
NMO Response Factor (area units/ppm C)
Results
Measured Concentration for Sample Tank, Ct (ppra NMO)
Measured Concentration for Condensate Trap, C (ppm C)
Noncondensible Organic Concentration (tank), Ct (ppra C)
Condensible Organic Concentration (trap), Cc (ppm C)
Note: Prior approval of Adninstrator required for blank substraction
TGNMO Concentration, C (ppn C)
Flue Gas Flow Rate
Emission Rate (rag/h)
••From EPA Method testing.
Figure 6.2. Continued
*"0 t^ C/J
03 p3 ro
CD fl> ft
VO—J
(-» •
ON
O
o
o
-------
Section No. 3.17.6
Date May 31, 1991
Page 11
Table 6.1. ACTIVITY MATRIX FOR CALCULATION CHECKS
Characteristic
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Analysis data
form
All data and calcu-
tions are shown
Visually check
Complete the
missing data
Calculations
Difference between
check and original
calculations should
not exceed round-off
error
Repeat all calcula-
tions starting with
raw data for hand
calculations; check
all raw data input
for computer calcu-
lations; hand calcu-
late one sample per
test
Indicate errors
on calculation
form, Figure 6.1
-------
o
o
o
-------
Section No. 3.17-7
Date May 31, 1991
Page 1
7.0 MAINTENANCE
The normal use of emission testing equipment subjects it to corrosive gases,
extremes in temperature, vibration, and shock. Keeping the equipment in good
operating order over an extended period of time requires knowledge of the equipment
and a program of routine maintenance which is performed quarterly or after 2830 L
(100 ft3) of operation, whichever is greater. In addition to the quarterly main-
tenance, a yearly cleaning of pumps and metering systems is recommended. Main-
tenance procedures for the various components are summarized in Table 8.1 at the
end of the section. The following procedures are not required, but are recommended
to increase the reliability of the equipment.
7.1 Pump
Several types of pumps may be used to perform Method 25; the two most common
are the fiber vane pump with in-line oiler and the diaphragm pump. The fiber vane
pump requires a periodic check of the oiler jar. Its contents should be translu-
cent; the oil should be changed if not translucent. Use the oil specified by the
manufacturer. If none is specified, use SAE-10 nondetergent oil. Whenever a fiber
vane pump starts to run erratically or during the yearly disassembly, the head
should be removed and the fiber vanes changed. Erratic operation of a diaphragm
pump is normally due to either a bad diaphragm (causing leakage) or to malfunctions
of the valves, which should be cleaned annually by complete disassembly.
7.2 Rotameter
Rotameters should be disassembled and cleaned according to the manufacturer's
instructions using recommended cleaning fluids every 3 months or upon erratic
operation.
7.3 Manometer
The fluid in the manometers should be changed whenever there is discoloration
or visible matter in the fluid, and during the yearly disassembly.
7.^ Sampling Train
All remaining sampling train components should be visually checked every 3
months and completely disassembled and cleaned or replaced yearly. Many items,
such as quick disconnects, should be replaced whenever damaged rather than checked
periodically. Normally, the best procedure for maintenance in the field is to have
on hand another entire sampling system, including a pump, probe, U-tubes, filter
holders, sample tanks, and heated sample line rather than replacing individual
components. It is recommended that the U-tubes be filled with nitrogen after
burnout to reduce oxidation of the metal surface. The sample tanks should be clean
and dry when being stored between tests.
7.5 Gas Chroroatograph and NDIR
Maintenance activities and schedules for gas chromatographs and NDIRs are make
and model specific. It is recommended that the analyst consult the operator's
manual for instructions relative to maintenance practices and procedures.
-------
Section No. 3-17.7
Date May 31, 1991
Page 2
Table 7.1. ACTIVITY MATRIX FOR EQUIPMENT MAINTENANCE CHECKS
o
Apparatus
Fiber vane
pump
Diaphragm
pump
Rotameter
Manometer
Sampling
train
components
Gas chroma-
tograph and
NDIR
Acceptance limits
In-line oiler
free of leaks
Leak- free valves
functioning properly
Clean and no errat-
ic behavior
No discoloration or
visible matter in
the fluid
No damage
See owner's manual
Frequency and method
of measurement
Periodically check
oiler jar; remove
head and change fiber
vanes
Clean valves during
yearly disassembly
Clean every 3 mo. or
whenever ball does
not move freely
Check periodically
and during disassem-
bly
Visually check every
3 mo.; completely
disassemble and
clean or replace
yearly
See owner's manual
Action if require-
ments are not met
Replace as
needed
Replace when
leaking or mal-
functioning
Replace
Replace parts
as needed
If failure noted,
replace appro-
priate components
See owner's manual
O
o
-------
Section No. 3-17-8
Date May 31, 1991
Page 1
8.0 AUDITING PROCEDURES
An audit is an independent assessment of data quality. Independence is
achieved if the individual(s) performing the audit and their standards and
equipment are different from the regular field team and their standards and
equipment. Routine quality assurance checks by a field team are necessary to
generate good quality data, but they are not part of the auditing procedure. Table
8.1 at the end of this section summarizes the quality assurance functions for
auditing.
Based on the requirements of Method 25 and the results of collaborative test-
ing of other EPA Test Methods, one performance audit is required when testing for
compliance for Standards of New Source Performance (and as required by other
government agencies) and is recommended when testing for other purposes; and a
second performance audit is recommended. The two performance audits are:
1. An audit of the sampling and analysis of Method 25 is required for NSPS
and recommended for other purposes.
2. An audit of the data processing is recommended.
It is suggested that a systems audit be conducted as specified by the quality
assurance coordinator in addition to these performance audits. The two performance
audits and the systems audit are described in detail in Subsections 8.1 and 8.2,
respectively.
8.1 Performance Audits
Performance audits are conducted to evaluate quantitatively the quality of
data produced by the total measurement system (sample collection, sample analysis,
and data processing). It is required that a cylinder gas performance audit be
performed once during every NSPS test utilizing Method 25 and it is recommended
that a cylinder gas audit be performed once during any enforcement source test
utilizing Method 25 conducted under regulations other than NSPS.
8.1.1 Performance Audit of the Field Test - As stated in Section 4.5 of Method 25
(40 CFR 60, Appendix A) and the "Instructions for the Sampling and Analysis of
Total Gaseous Nonmethane Organics from Quality Assurance Audit Cylinders using EPA
Method 25 Procedures" (supplied with the EPA audit gas cylinders), a set of two
audit samples are to be collected in the field (not laboratory) from two different
concentration gas cylinders at the same time the compliance test samples are being
collected. The two audit samples are then analyzed concurrently and in exactly the
same manner as the compliance samples to evaluate the tester's and analyst's
technique and the instrument calibration. The information required to document the
collection and analysis of the audit samples has been included on the example data
sheets shown later in Figures 8.1 and 8.3- The audit analyses shall agree within
20 percent of the actual cylinder concentrations.
The operator of the affected facility is responsible for informing local,
state, and federal agencies of the test program and of the test schedule.
Therefore, the operator of the affected facility is responsible for requesting (and
handling of) the audit samples from the agency responsible for observing the
compliance test. The tester is responsible for informing his client of the details
of any method used and any audit samples required to validate the compliance test.
These details are usually part of a pretest meeting, but many times pretest
meetings are not necessary depending on the scope of the work and permit
requirements.
-------
Section No. 3.17-8
Date May 31, 1991
Page 2 • x—v
The facility (or tester) may obtain audit cylinders by contacting the agency —
responsible for observing and/or evaluating the compliance test and informing the
agency the time and location of the compliance test. This should be done at least
l5 days prior to the test date. The responsible agency will contact: U.S.
Environmental Protection Agency, Atmospheric Research and Exposure Laboratory,
Quality Assurance Division (MD-77), Research Triangle Park, North Carolina 277H
and have the cylinders shipped to the specified site.
Responsibilities of the Audit Supervisor - The primary responsibilities of
the audit supervisor are to ensure that the proper audit gas cylinders are ordered
and safe-guarded, and to interpret the results obtained by the analyst.
When notified by the testing company 'that a test is to be conducted, the audit
supervisor will order the proper cylinders from the EPA's Quality Assurance
Division. Generally the audit cylinders will be shipped (at EPA's expense)
directly to the specified site. However, if the audit supervisor will be on-site
during the compliance test, the audit cylinders may be shipped to the testing
company for transport to the sampling site. Since the audit cylinders are sealed
by EPA, the testing firm will not be allowed to collect any audit gas without
breaking the seal. The audit gas concentration^) should be in the range of 50/»
below to 100% above the applicable standard. If two cylinders are not available,
then one cylinder can be used.
The audit supervisor must ensure that the audit gas cylinder(s) are shipped to
the correct address, and to prevent vandalism, verify that they are stored in a
safe location both before and after the audit. Also, the audit cylinders should
not be analyzed when the pressure drops below 200 psi. The audit1 supervisor then
ensures that the audits are conducted as described below. At the conclusion of the
collection of the audit samples, the cylinders are then returned to shipping
laboratory as per the instructions supplied with the cylinders at the expense of
the facility. If the tester is to transport the audit cylinders to his home
laboratory for shipment back to the EPA/QAD contractor, the audit supervisor will
seal both cylinders to ensure that additional audit sample gas cannot be collected
without breaking the seal.
The audit supervisor must also interpret the audit results. Indication of
acceptable results may be obtained immediately by reporting the audit compliance
test results in ppm by telephone to the responsible enforcement agency. The tester
must also include the results of both audit samples, their identification numbers,
and the analyst's name along with the results of the compliance determination
samples in the appropriate reports to the EPA regional office or the appropriate
enforcement agency during the 30-day period.
When the measured concentration agrees within 20% of the true value, the audit
results are considered acceptable. Failure to meet the 20-percent specification
may require reanalysis of the audit samples and compliance test samples,
reauditing, or retests until the audit problems are resolved. However, if the
audit results do not affect the compliance or noncompliance status of the affected
facility, the Administrator (enforcement agency) may waive the reanalysis, further
audits, or retest requirements and accept the results of the compliance test. For
example, if the audit results average 38.6% low, the compliance results would be
divided by (1 - 0.386) to determine the correlated effect. If the audit results
average 58.3% high, the compliance sample results would be divided by (1 + 0.583)
to determine the effect. When the compliance status of the source is the same with
and without the correlated value, then the responsible agency may accept the
results of the test. While steps are being taken to resolve audit analysis
problems, the Administrator may also choose to use the data to determine the
o
o
-------
Section No. 3-17.8
Date May 31, 1991
Page 3
compliance or noncompliance of the affected facility.
The same analysts, analytical reagents, and analytical system shall be used
for the compliance samples and the EPA audit samples; if this condition is met, and
the same testing firm is collecting other sets of compliance samples, auditing of
subsequent compliance analyses for the same enforcement agency within 30 days is
not required. An audit sample set may not be used to validate different sets of
compliance samples under the jurisdiction of different enforcement agencies, unless
prior arrangements are made with both enforcement agencies.
During the audit, the audit supervisor should record the coded cylinder
number(s) and cylinder pressure(s) on the "Audit Report" form, Figure 8.1. The
individual being audited must not be told the actual audit concentrations or the
calculated audit percent accuracy.
On-site Collection of Audit Sample(s) - The cylinder gas performance audit
must be conducted in the field (not laboratory) at the same time the compliance
test samples are being taken. A maximum of 5L of audit gas is to be used for each
test run unless multiple tanks are required for sampling. The tester will be
required to supply a two-stage regulator (CGA - 350), a glass manifold or Teflon
tee connection and other suitable Swagelok fittings (they are not supplied) for use
with the audit gas cylinder. The recommended procedures for conducting the on-site
audit sample collection are as follows:
1. The audit supervisor (agency representative) should verify that the seal
affixed by shipping or supplying laboratory is still intact. After the
seal has been checked by the audit supervisor, the tester may break the
seal. However, if the audit supervisor is not present at the time of the
audit, the tester may break the seal and proceed with the audit.
2. The tester should set up the Method 25 sampling train and perform the leak
check.
3. The audit gas from the cylinder has to be sampled at atmospheric pressure
either from a glass manifold or through a Teflon tee connection. This can
be done by attaching both the cylinder and the probe of the Method 25
sampling train to two of the manifold or tee connections while excess gas
flows out through the remaining connection as shown in Figure 8.2. This
can be accomplished by starting the cylinder gas flow into the manifold or
tee with the sampling train flow turned off. Then, turn on the sampling
train flow while adjusting the flow from the audit cylinder to ensure
excess audit gas flow from the manifold or tee. After the proper sampling
flow rate has been obtained in the sampling train, adjust the audit cylin-
der so only a few cubic centimeters of excess gas is discharged from the
manifold or tee. The tester must ensure that the audit gas is conserved.
4. Use the same sampling flow rate and sample volume as used for field test
samples. When a constant flow rate can no longer be maintained by the
sampling train, it should be turned off and then the audit cylinder shut
off. Ensure that the audit cylinder is closed tight to prevent leakage.
If the compliance test requires more than one sample tank to complete a
run, the audit should use the same number of tanks required by the average
run.
5. The same procedures are repeated for the second audit cylinder using"a
separate sampling train.
6. The sampling trains containing the audit samples should be stored and
shipped in the same manner as and along with the field test samples.
7. In all cases, it is recommended that the audit supervisor reseal the audit
cylinders to ensure no tampering. However, if the test firm is to return
the cylinder to shipping or supplying laboratory, it is mandatory that the
-------
o
Section No. 3.17-8
Date May 31, 1991
Page 4
audit cylinders are resealed by the audit supervisor.
8. The audit cylinders are to be returned immediately after the test to the
EPA/QAD contractor at the cost of the facility (or tester if applicable)
either by ground transportation or air cargo. They are not to be shipped
collect.
Analysts of Audit Sample(s) - Analyze the collected audit sample fractions
(condensibles and noncondensibles) at the same time as the Method 25 compliance
test samples. Follow the procedures described in the method for sample analysis,
calibration, and calculations. The same analysts, analytical reagents, and
analytical system shall be used,for b9th the compliance test samples and the EPA
audit samples.
Reporting of Audit Sample(s) Results - The reporting of the audit results
should be the responsibility of the tester to ensure taht the data is acceptable
and valid. The audit sample results are to be reported to the responsible agency
by the testing firm in terms of condensibles (U-trap fraction), noncondensibles
(tank fraction), and total (sum of both fractions) as parts-per-million carbon (ppm
C). The agency will in turn report the results to the EPA/QAD contractor for
continuing evaluation of the Method 25 audit program. Additionally, the tester
must supply document in the test report, the results of both audit samples as
described above, their identification numbers, and the analyst's name along with
the results of the compliance determination samples. The operator of the affected
facility is responsible for the dissemination of any compliance sample results and
the correct distribution of reports to the EPA regional office or the appropriate
enforcement agency during the 30-day period for which the audit samples represent. (i
8.1.2 Performance Audit of Data Processing - Calculation errors are prevalent in
processing data. Data processing errors can be determined by auditing the recorded
data on the field and laboratory forms. The original and audit (check) calcula-
tions should agree within round-off error; if not, all of the remaining data should
be checked. The data processing may also be audited by providing the testing
laboratory with specific data sets (exactly as would appear in the field), and by
requesting that the data calculation be completed and that the results be returned
to the agency. This audit is useful in checking both computer programs and manual
methods of data processing.
8.2 Systems Audit
A systems audit is an on-site, qualitative inspection and review of the total
measurement system (sample collection, sample analysis, etc.). Initially, a
systems audit is recommended for each enforcement source test, defined here as a
series of three runs at one source. After the test team gains experience with the
method, the frequency of auditing may be reduced — for example, to once every four
tests.
The auditor should have extensive background experience in source sampling,
specifically with the measurement system being audited. The functions of the
auditor are summarized below:
1. Inform the testing team of the results from previous audits, specifying
any area(s) that need special attention or improvement.
2. Observe procedures and techniques of the field team during sample collec-
tion.
3. Check/verify records of apparatus calibration checks and quality control
used in the laboratory analysis of control samples from previous source
-------
Section No. 3.1?.8
Date May 31, 1991
Page 5
tests, where applicable.
4. Record the results of the audit, and forward them with comments to the
test team management so that appropriate corrective action may be
initiated.
While on site, the auditor observes the source test team's overall perfor-
mance, including the following specific operations:
1. Setting up and leak testing the sampling train.
2. Collecting the sample at a constant rate at the specified flow rate.
3. Conducting the final leak check and recovery of the samples.
4. Sample documentation procedures, sample recovery, and preparation of
samples for shipment.
Figure 8.3 is a suggested checklist for the auditor.
-------
Section No. 3.17-8
Date Hay 31, 1991
Page 6
AUDIT REPORT
O
Part A. - To be filled out by organization supplying audit cylinders.
1. Organization supplying audit sample(s) and shipping address
2. Audit supervisor, organization, and phone number
3. Shipping instructions: Name, Address, Attention
4. Guaranteed arrival date for cylinders - '
5. Planned shipping date for cylinders -
6. Details on audit cylinders from last analysis
d. Audit gas (es) /balance gas..
f . Cylinder construction. .....
Low cone.
N2
Aluminum
High cone
N2
Aluminum
Part B. - To be filled out by audit supervisor.
1. Process sampled
Audit location
2.
3-
Name of individual audit
Audit date
O
5. Audit cylinders sealed
6. Audit results:
d. Measured concentration, ppm C
e. Actual audit concentration, ppm C
f. Audit accuracy:1
Percent accuracy1 =
Measured Cone. - Actual Cone. x 100
Actual Cone.
g. Problems detected (if any)
Low
cone.
cylinder
High
cone.
cylinder
1The audit accuracy is calculated on the total concentration only.
Figure 8.1. Field audit report form.
O
-------
Section No. 3.17.8
Date May 31, 1991
Page 7
MANOMETER
EXCESS
FLOW
FLOW
CONTROL
VALVE
ROTAMETER
ROTAMETER
TEFLON TEE"
OR MANIFOLD
4-
SAMPLE
TANK
VALVE
CONDENSATE
TRAP
SAMPLE
TANK
AUDIT
CYLINDER
(in
>|Lj
Figure 8.2. Schematic of Method 25 audit system.
-------
Section No. 3.17-8
Date May 31. 1991
Page 8
o
Yes
1 v
No
Comments
Operation
PRESAMPLING PREPARATION
1. Knowledge of process operations
2. Calibration of pertinent equipment, in
particular, temperature readouts and flowmeters
3. Selection and checkout of equipment for proper
sampling techniques
ON-SITE MEASUREMENTS
* 4. Sampling system properly assembled
5. ^Sampling system leak check acceptable
6. Sample probe and filter at proper temperature
7. Sample system purged properly
8. Constant rate sampling properly conducted
9. Heater systems maintained at proper temperatures
10. Proper number of samples & sampling time
11. Recording of pertinent process conditions during
sample collection, samples properly identified,
and calculations properly conducted
POSTSAMPLING
12. Results of audit (+ 20% or other value)
13. Oxidation catalyst efficiency test acceptable
14. Reduction catalyst efficiency test acceptable
15. NMO linearity and calibration test acceptable
16. C02 linearity and calibration test acceptable
17. NMO analyzer performance test acceptable
18. Condensible organic recovery system leak check
19. System background test acceptable
20. Temperature, volumes, pressures, and concen-
trations properly recorded
21. Analytical results properly calculated
COMMENTS
O
Figure 8.3. Method 25 checklist to be used by auditors.
O
-------
Section No. 3.17-8
Date May 31, 1991
Page 9
Table 8.1. ACTIVITY MATRIX FOR AUDITING PROCEDURES
Apparatus
Performance
audit of
analytical phase
Acceptance limits
Measured relative
error of audit
samples less than
20% for both samples
Frequency and method
of measurement
Frequency: Once during
every enforcement
source test*
Method; Measure audit
samples and compare
results to true values
Action if
requirements
are not met
Review operating
technique and
repeat audit,
repeat test,
reject test, or
accept results
Data processing
errors
Original and checked
calculations agree
within round-off
error
Frequency: Once during
every enforcement
source test*
Method; Independent
calculations starting
with recorded data
Check and correct
all data from the
audit period
represented by
the checked data
Systems audit—
observance
of technique
Operational tech-
nique as described
in this section of
the Handbook
Frequency; Once during
every enforcement
source test* until
experience gained,
then every fourth
test
Method; Observation of
techniques assisted
by audit checklist,
Figure 8.3
Explain to test
team their devia-
tions from rec-
commended tech-
niques and note
on Fig. 8.3
*As defined here, a source test for enforcement of the NSPS comprises a series of
runs at one source. Source tests for purposes other than enforcement of NSPS may
be audited at the frequency determined by the applicable group.
-------
o
o
o
-------
Section No. 3.17-9
Date May 31, 1991
Page 1
9.0 RECOMMENDED STANDARDS FOR ESTABLISHING TRACEABILITY
To achieve data of desired quality, two essential considerations are
necessary: (1) the measurement process must be in a state of statistical control
at the time of the measurement, and (2) the systematic errors, when combined with
the random variation (errors or measurment), must result in an acceptable
uncertainty. As evidence in support of good quality data, it is necessary to
perform quality control checks and independent audits of the measurement process;
to document these data; and to use materials, instruments, and measurement
procedures that can be traced to an apropriate standard of reference.
Data must be routinely obtained by repeat measurements of standard reference
samples (primary, secondary, and/or working standards) and the establishment of a
condition of process control. The working calibration standards should be
traceable to standards of higher accuracy. It is recommended, but not required,
that the analyst use a NBS-SRM for propane to make the analysis traceable to an NBS
Standard Reference Material.
Audit samples (as discussed in Section 3.17.8) must be used to validate test
results for compliance determination purposes and are recommendeed as an
independent check on the measurement process when the method is performed for other
purposes. This makes all the compliance determination samples traceable to the
same audit source at EPA.
-------
o
o
o
-------
Section No. 3.17.10
Date May 31, 1991
Page 1
10.0 REFERENCE METHOD
Method 25 - Determination of Total Gaseous NonMethane
Organic Emissions as Carbon
1. Applicability and Principle
1.1 Applicability. This method applies to the measurement of volatile organic
compounds (VOC) as total gaseous nonmethane organics (TGNMO) as carbon in source
emissions. Organic particulate matter will interfere with the analysis, and,
therefore, a particulate filter is required. The minimum detectable concentration
for the method is 50 ppm as carbon.
When carbon dioxide (C02) and water vapor are present together in the stack,
they can produce a positive bias in the sample. The magnitude of the bias depends
on the concentration of C02 and water vapor. As a guideline, multiply the C02
concentration, expressed as volume percent, times the water vapor concentration.
If this product does not exceed 100, the bias can be considered insignificant. For
example, the bias is not significant for a source having 10 percent C02 and 10
percent water vapor, but it would be significant for a source near the detection
limit having 10 percent C02 and 20 percent water vapor.
This method is not the only method that applies to the measurement of TGNMO.
Costs, logistics, and other practicalities of source testing may make other test
methods more desirable for measuring VOC contents of certain effluent streams.
Proper judgment is required in determining the most applicable VOC test method.
For example, depending upon the molecular weight of the organics in the effluent
stream, a totally automated semicontinuous nonmethane organics (NMO) analyzer
interfaced directly to the source may yield accurate results. This approach has
the advantage of providing emission data semicontinuously over an extended time
period.
Direct measurement of an effluent with a flame ionization detector (FID)
analyzer may be appropriate with prior characterization of the gas stream and
knowledge that the detector responds predictably to the organic compounds in the
stream. If present, methane (CHA) will, of course, also be measured. The FID can
be applied to the determination of the mass concentration of the total molecular
structure of the organic emissions under any of the following limited conditions:
(1) where only one compound is known to exist; (2) when the organic compounds
consist of only hydrogen and carbon; (3) where the relative percentages of the
compounds are known or can be determined, and the FID responses to the compounds
are known; (4) where a consistent mixture of the compounds exists before and after
emission control and only the relative concentrations are to be assessed; or (5)
where the FID can be calibrated against mass standards of the compounds emitted
(solvent emissions, for example).
Another example of the use of a direct FID is as a screening method. If there
is enough information available to provide a rough estimate of the analyzer
accuracy, the FID analyzer can be used to determine the VOC content of an
uncharacterized gas stream. With a sufficient buffer to account for possible
inaccuracies, the direct FID can be a useful tool to obtain the desired results
without costly exact determination. In situations where a qualitative/
quantitative analysis of an effluent stream is desired or required, a gas
chromatographic FID system may apply. However, for sources emitting numerous
organics, the time and expense of this approach will be formidable.
1
-------
Section No. 3.17.10
Date May 31. 1991
Page 2
o
1.2 Principle. An emission sample is withdrawn from the stack at a constant
rate through a heated filter and a chilled condensate trap by means of an evacuated
sample tank. After sampling is completed, the TGNMO are determined by
independently analyzing the condensate trap and sample tank fractions and combining
the analytical results. The organic content of the condensate trap fraction is
determined by oxidizing the NMO to C02 and quantitatively collecting the effluent
in an evacuated vessel; then a portion of the C02 is reduced to CHA and measured by
an FID. The organic content of the sample tank fraction is measured by injecting a
portion of the sample into a gas chromatographic column to separate the NMO from
carbon monoxide (CO). C02. and CH/,; the NMO are oxidized to C02. reduced to CHA.
and measured by an FID. In this manner, the variable response of the FID
associated with different types of organics is eliminated.
2. Apparatus
2.1 Sampling. The sampling system consists of a heated probe, heated filter,
condensate trap, flow control system, and sample tank (Figure 25~1). The TGNMO
sampling equipment can be constructed from commercially available components and
components fabricated in a machine shop. The following equipment is required:
2.1.1 Heated Probe. 6.4-mm (1/4-in.) OD stainless steel tubing with a heating
system capable of maintaining a gas temperature at the exit end of at least 129°C
(265°F). The probe shall be equipped with a thermocouple at the exit end to
monitor the gas temperature. x""N.
A suitable probe is shown in Figure 25~1• The nozzle is an elbow f ittingf A
attached to the front end of the probe while the thermocouple is inserted in the\__y
side arm of a tee fitting attached to the rear of the probe. The probe is wrapped
with a suitable length of high temperature heating tape, and then covered with two
layers of glass cloth insulation and one layer of aluminum foil.
NOTE; If it is not possible to use a heating system for safety reasons, an
unheated system with an in-stack filter is a suitable alternative.
2.1.2 Filter Holder. 25-mm (15/l6-in.) ID Gelman filter holder with stainless
steel body and stainless steel support screen with the Viton 0-ring replaced by a
Teflon 0-ring.
NOTE; Mention of trade names or specific products does not constitute
endorsement by the Environmental Protection Agency.
2.1.3 Filter Heating System. A metal box consisting of an inner and an outer
shell separated by insulating material with a heating element in the inner shell
capable of maintaining a gas temperature at the filter of 121 +_ 3°C (250 +_ 5°F).
A suitable heating box is shown in Figure 25-2. The outer shell is a metal
box that measures 102 mm x 280 mm x 292 mm (4 in. x 11 in. x 11 1/2 in.), while the
inner shell is a metal box measuring 76 mm x 229 mm x 2*41 mm (3 in. x 9 in. x 9 1/2
in.). The inner box is supported by 13-mm (1/2-in.) phenolic rods. The void space
between the boxes is filled with fiberfrax insulation which is sealed in place by
means of a silicon rubber bead around the upper sides of the box. A removable lid
made in a similar manner, with a 25-mm (1-in.) gap between the parts, is used to
cover the heating chamber.
The inner box is heated with a 250-watt cartridge heater, shielded by a f~\
stainless steel shroud. The heater is regulated by a thermostatic temperaturef
controller which is set to maintain a temperature of 121° C as measured by a
thermocouple in the gas line just before the filter. An additional thermocouple is
-------
Section No. 3.17.10
Date May 31, 1991
Page 3
used to monitor the temperature of the gas behind the filter.
2.1.4 Condensate Trap. 9.5-mm (3/8-in.) OD 316 stainless steel tubing bent
into a U-shape. Exact dimensions are shown in Figure 25-3. The tubing shall be
packed with coarse quartz wool, to a density of approximately 0.11 g/cc before
bending. While the condensate trap is packed with dry ice in the Dewar, an ice
bridge may form between the arms of the condensate trap making it difficult to
remove the condensate trap. This problem can be prevented by attaching a steel
plate between the arms of the condensate trap in the same plane as the arms to
completely fill the intervening space.
2.1.5 Valve. Stainless steel shut-off valve for starting and stopping sample
flow.
2.1.6 Metering Valve. Stainless steel control valve for regulating the sample
flow rate through the sample train.
2.1.7 Rotameter. Glass tube with stainless steel fittings, capable of
measuring sample flow in the range of 60 to 100 cc/min.
2.1.8 Sample Tank. Stainless steel or aluminum tank with a minimum volume of
4 liters.
2.1.9 Mercury Manometer or Absolute Pressure Gauge. Capable of measuring
pressure to within 1 mm Hg in the range of 0 to 900 mm.
2.1.10 Vacuum Pump. Capable of evacuating to an absolute pressure of 10 mm
Hg.
2.2. Condensate Recovery Apparatus. The system for the recovery of the
organics captured in the condensate trap consists of a heat source, oxidation
catalyst, nondispersive infrared (NDIR) analyzer and an intermediate collection
vessel (ICV). Figure 25-4 is a schematic of a typical system. The system shall be
capable of proper oxidation and recovery, as specified in Section 5«1. The
following major components are required:
2.2.1 Heat Source. Sufficient to heat the condensate trap (including
connecting tubing) to a temperature of 200°C. A system using both a heat gun and
an electric tube furnace is recommended.
2.2.2 Heat Tape. Sufficient to heat the connecting tubing between the water
trap and the oxidation catalyst to 100°C.
2.2.3. Oxidation Catalyst. A suitable length of 9.5-mm (3/8-in.) OD Inconel
600 tubing packed with 15 cm (6 in.) of 3.2-mm (1/8-in.) diameter 19 percent
chromia on alumina pellets. The catalyst material is packed in the center of the
catalyst tube with quartz wool packed on either end to hold it in place. The
catalyst tube shall be mounted vertically in a 650°C tube furnace.
2.2.4 Water Trap. Leak proof, capable of removing moisture from the gas
stream.
2.2.5 Syringe Port. A 6.4-mm (1/4-in.) OD stainless steel tee fitting with a
-------
Section No. 3.17.10
Date May 31, 1991
Page 4
rubber septum placed in the side arm.
2.2.6 NDIR Detector. Capable of indicating C02 concentration in the range of
zero to 5 percent, to monitor the progress of combustion of the organic compounds
from the condensate trap.
2.2.7 Flow-Control Valve. Stainless steel, to maintain the trap conditioning
system near atmospheric pressure.
2.2.8 Intermediate Collection Vessel. Stainless steel or aluminum, equipped
with a female quick connect. Tanks with nominal volumes of at least 6 liters are
recommended. \ r
2.2.9 Mercury Manometer or Absolute Pressure Gauge.
pressure to within 1 mm Hg in the range of 0 to 900 mm.
Capable of measuring
2.2.10 Syringe. 10-ml gas-tight, glass syringe equipped with an appropriate
needle. •""•'
2.3 NMO Analyzer. The NMO analyzer is a gas chromatograph (GC) with
backflush capability for NMO analysis and is equipped with an oxidation catalyst,
reduction catalyst, and FID. Figures 25-5 and 25-6 are schematics of a typical NMO
analyzer. This semicontinuous GC/FID analyzer shall be capable of: (1) separating
CO, C02 , and CH4 from NMO, (2) reducing the C02 to CHA and quantifying as CH^,. and
(3) oxidizing the NMO to C02, reducing the C02 to CH4 and quantifying as CH4.
according to Section 5-2. The analyzer consists of the following major components:
2.3.1 Oxidation Catalyst. A suitable length of 9.5-mm (3/8-in.) OD Inconel
600 tubing packed with 5-1 cm (2 in.) of 19 percent chromia on 3«2-mm (1/8-in.)
alumina pellets. The catalyst material is packed in the center of the tube
supported on either side by quartz wool. The catalyst tube must be mounted
vertically in a 650°C furnace.
2.3.2 Reduction Catalyst. A 7.6-cm (3-in.) length of 6.4-mm (1/4-in.) OD
Inconel tubing fully packed with 100-mesh pure nickel powder. The catalyst tube
must be mounted vertically in a 400°C furnace.
2.3.3 Separation Column(s). A 30-cm (1-ft) length of 3.2-mm (1/8-in.) OD
stainless steel tubing packed with 60/80 mesh Unibeads IS followed by a 6l-cm (2-
ft) length of 3.2-mm (1/8-in.) OD stainless steel tubing packed with 60/80 mesh
Carbosieve G. The Carbosieve and Unibeads columns must be baked separately at
200°C with carrier gas flowing through them for 24 hours before initial use.
f
2.3.4 Sample Injection System. A 10-port GC sample injection valve fitted
with a sample loop properly sized to interface with the NMO analyzer (1-cc loop
recommended).
2.3.5 FID. An FID meeting the following specifications is required:
2.3.5-1 Linearity. A linear response (j+5 percent) over the operating range
as demonstrated by the procedures established in Section 5.2.3.
2.3.5.2 Range. A full scale range of 10 to 50,000 ppm CHA. Signal
o
o
-------
Section No. 3.17.10
Date May 31, 1991
Page 5
attenuators shall be available to produce a minimum signal response of 10 percent
of full scale.
2.3.6 Data Recording System. Analog strip chart recorder or digital
integration system compatible with the FID for permanently recording the
analytical results.
2.4 Other Analysis Apparatus.
2.4.1 Barometer. Mercury, aneroid, or other barometer capable of measuring
atmospheric pressure to within 1 mm Hg.
2.4.2 Thermometer. Capable of measuring the laboratory temperature to within
1°C.
2.4.3 Vacuum Pump. Capable of evacuating to an absolute pressure of 10 mm
Hg.
2.4.4 Syringes. 10-ul and 50-ul liquid injection syringes.
2.4.5 Liquid Sample Injection Unit. 316 SS U-tube fitted with an injection
septum, see Figure 25~7-
3. Reagents
3.1 Sampling. The following are required for sampling:
3-1.1 Crushed Dry Ice.
3-1.2 Coarse Quartz Wool. 8 to 15 urn.
- ' .•'
3.1.3 Filters. Glass fiber filters, without organic binder.
3.2 NMO Analysis. The following gases are needed:
3.2.1 Carrier Gases. Zero grade helium (He) and oxygen (02) containing less
than 1 ppm C02 and less than 0.1 ppm C as hydrocarbon.
3-2.2 Fuel Gas. Zero grade hydrogen (H2), 99.999 percent pure.
3.2.3 Combustion Gas. Zero grade air or 02 as required by the detector.
3-3 Condensate Analysis. The following gases are needed:
3.3-1 Carrier Gas. Zero grade air, containing less than 1 ppm C.
3.3«2 Auxiliary 02. Zero grade 02, containing less than 1 ppm C.
3.3.3 Hexane. ACS grade, for liquid injection.
3.3.4 Decane. ACS grade, for liquid injection.
3.4 Calibration. For all calibration gases, the manufacturer must recommend
-------
Section No. 3-17-10
Date May 31, 1991 -^
Page 6 (l
a maximum shelf life for each cylinder (i.e., the length of time the gas
concentration is not expected to change more than +_ 5 percent from its certified
value). The date of gas cylinder preparation, certified organic concentration, and
recommended maximum shelf life must be affixed to each cylinder before shipment
from the gas manufacturer to the buyer. The following calibration gases are
required:
3.4.1 Oxidation Catalyst Efficiency Check Calibration Gas. Gas mixture
standard with nominal concentration of 1 percent methane in air.
3.4.2 FID Linearity and NMO Calibration Gases. Three gas mixture standards
with nominal propane concentrations of 20 ppm, 200 ppm, and 3000 ppm, in air.
3.4.3 C02 Calibration Gases. Three gas mixture standards with nominal C02
concentrations of 50 ppm, 500 ppm, and 1 percent, in air.
NOTE; Total NMO of less than 1 ppm required for 1 percent mixture.
3.4.4 NMO Analyzer System Check Calibration Gases. Four calibration gases
are needed as follows:
3-4.4.1 Propane Mixture. Gas mixture standard containing (nominal) 50 ppm
CO, 50 ppm CH4, 2 percent C02, and 20 ppm C3H8, prepared in air.
3.4.4.2 Hexane. Gas mixture standard containing (nominal) 50 ppm hexane in s~*\
o
3.4.4.3 Toluene. Gas mixture standard containing (nominal) 20 ppm toluene in
air.
3.4.4.4 Methanol. Gas mixture standard containing (nominal) 100 ppm methanol
in air.
4. Procedure
4.1 Sampling.
4.1.1 Cleaning Sampling Equipment. Before its initial use and after each
subsequent use, a condensate trap should be thoroughly cleaned and checked to
insure that it is not contaminated. Both cleaning and checking can be accomplished
by installing the trap in the condensate recovery system and treating it as if it
were a sample. The trap should be heated as described in the final paragraph of
Section 4.3.3- A trap may be considered clean when the C02 concentration in its
effluent gas drops below 10 ppm. This check is optional for traps that have been
used to collect samples which were then recovered according to the procedure in
Section 4.3.3.
4.1.2 Sample Tank Evacuation and Leak Check. Evacuate the sample tank to 10
mm Hg absolute pressure or less. Then close the sample tank valve, and allow the
tank to sit for 60 minutes. The tank is acceptable if no change in tank vacuum is
noted. The evacuation and leak check may be conducted either in the laboratory or
the field. The results of the leak check should be included in the test report.
4.1.3 Sample Train Assembly. Just before assembly, measure the tank vacuum
using a mercury U-tube manometer or absolute pressure gauge. Record this vacuum,
V.
O
-------
Section No. 3.17.10
Date May 31, 1991
Page 7
the ambient temperature, and the barometric pressure at this time. Close the
sample tank valve and assemble the sampling system as shown in Figure 25~1.
Immerse the condensate trap body in dry ice. The point where the inlet tube joins
the trap body should be 2.5 to 5 cm above the top of the dry ice.
4.1.4 Pretest Leak Check. A pretest leak check is required. Calculate or
measure the approximate volume of the sampling train from the probe trip to the
sample tank valve. After assembling the sampling train, plug the probe tip, and
make certain that the sample tank valve is closed. Turn on the vacuum pump, and
evacuate the sampling system from the probe tip to the sample tank valve to an
absolute pressure of 10 ppm Hg or less. Close the purge valve, turn off the pump,
wait a minimum period of 5 minutes, and recheck the indicated vacuum. Calculate
the maximum allowable pressure change based on a leak rate of 1 percent of the
sampling rate using Equation 25-1, Section 6.2. If the measured pressure change
exceeds the calculated limit, correct the problem before beginning sampling. The
results of the leak check should be included in the test report.
4.1.5 Sample Train Operation. Unplug the probe tip, and place the probe into
the stack such that the probe is perpendicular to the duct or stack axis; locate
the probe tip at a single preselected point of average velocity facing away from
the direction of gas flow. For stacks having a negative static pressure, seal the
El sample port sufficiently to prevent air in-leakage around the probe. Set the probe
temperature controller to 129°C (265°F) and the filter temperature controller to
121°C (250°F). Allow the probe and filter to heat for about 30 minutes before
purging the sample train.
Close the sample valve, open the purge valve, and start the vacuum pump. Set
the flow rate between 60 and 100 cc/min, and purge the train with stack gas for at
least 10 minutes. When the temperatures at the exit ends of the probe and filter
are within their specified range, sampling may begin.
Check the dry ice level around the condensate trap, and add dry ice if
necessary. Record the clock time. To begin sampling, close the purge valve and
stop the pump. Open the sample valve and the sample tank valve. Using the flow
control valve, set the flow through the sample train to the proper rate. Adjust
the flow rate as necessary to maintain a constant rate (+^10 percent) throughout the
duration of the sampling period. Record the sample tank vacuum and flowmeter
setting at 5~m±nute intervals. (See Figure 25-8.) Select a total sample time
greater than or equal to the minimum sampling time specified in the applicable
subpart of the regulation; end the sampling when this time period is reached or
when a constant flow rate can no longer be maintained because of reduced sample
tank vacuum.
NOTE; If sampling had to be stopped before obtaining the minimum sampling
time (specified in the applicable subpart) because a constant flow rate could not
be maintained, proceed as follows: After closing the sample tank valve, remove the
used sample tank from the sampling train (without disconnecting other portions of
the sampling train). Take another evacuated and leak-checked sample tank, measure
and record the tank vacuum, and attach the new tank to the sampling train. After
the new tank is attached to the sample train, proceed with the sampling until the
required minimum sampling time has been exceeded.
4.2 Sample Recovery. After sampling is completed, close the flow control
valve, and record the final tank vacuum; then record the tank temperature and
barometric pressure. Close the sample tank valve, and disconnect the sample tank
from the sample system. Disconnect the condensate trap at the flowmetering system,
-------
Section No. 3.17.10
Date Hay 31, 1991
Page 8 S~\
and tightly seal both ends of the condensate trap. Do not Include the probe from
the stack to the filter as part of the condensate sample. Keep the trap packed in
dry ice until the samples are returned to the laboratory for analysis. Ensure that
the test run number is properly identified on the condensate trap and the
sample tank(s).
4.3 Condensate Recovery. See Figure 25-9. Set the carrier gas flow rate,
and heat the catalyst to its operating temperature to condition the apparatus.
4.3.1 Daily Performance Checks. Each day before analyzing any samples,
perform the following tests:
*. J
4.3.1.1 Leak Check, With the carrier gas inlets and the flow control valve
closed, install a clean condensate trap in the system, and evacuate the system to
10 mm Hg absolute pressure or less. Close the vacuum pump valve and turn off the
vacuum pump. Monitor the system pressure for 10 minutes. The system is acceptable
if the pressure change is less than 2 mm Hg.
4.3.1.2 System Background Test. Adjust the carrier gas and auxiliary oxygen
flow rate to their normal values of 100 cc/min and 150 cc/min, respectively, with
the sample recovery valve in vent position. Using a 10-ml syringe withdraw a
sample from the system effluent through the syringe port. Inject this sample into
the NMO analyzer, and measure the C02 content. The system background is acceptable
if the C02 concentration is less than 10 ppm.
4.3.1.3 Oxidation Catalyst Efficiency Check. Conduct a catalyst efficiency \J
test as specified in Section 5.1-2 of this method. If the criterion of this test
cannot be met, make the necessary repairs to the system before proceeding.
4.3.2 Condensate Trap C02 Purge and Sample Tank Pressurization. After
sampling is completed, the condensate trap will contain condensed water and
organics and a small volume of sampled gas. This gas from the stack may contain a
significant amount of C02 which must be removed from the condensate trap before the
sample is recovered. This is accomplished by purging the condensate trap with zero
air and collecting the purged gas in the original sample tank.
Begin with the sample tank and condensate trap from the test run to be
analyzed. Set the four-port valve of the condensate recovery system in the C02
purge position as shown in Figure 25-9- With the sample tank valve closed, attach
the sample tank to the sample recovery system. With the sample recovery valve in
the vent position and the flow control valve fully open, evacuate the manometer or
pressure gauge to the vacuum of the sample tank. Next, close the vacuum pump
valve, open the sample tank valve, and record the tank pressure.
Attach the dry-ice-cooled condensate trap to the recovery system, and
initiate the purge by switching the sample recovery valve from vent to collect
position. Adjust the flow control valve to maintain atmospheric pressure in the
recovery system. Continue the purge until the C02 concentration of the trap
effluent is less than 5 ppm. C02 concentration in the trap effluent should be
measured by extracting syringe samples from the recovery system and analyzing the
samples with the NMO analyzer. This procedure should be used only after the NDIR
response has reached a minimum level. Using a 10-ml syringe, extract a sample from
the syringe port prior to the NDIR, and inject this sample into the NMO analyzer.
o
-------
Section No. 3.17.10
Date May 31, 1991
Page 9
After the completion of the C02 purge, use the carrier gas bypass valve to
pressurize the sample tank to approximately 1060 mm Hg absolute pressure with zero
air.
4.3.3 Recovery of the Condensate Trap Sample. See Figure 25-10. Attach the
ICV to the sample recovery system. With; the sample recovery valve in a closed
position, between vent and collect, and the flow control and ICV valves fully open,
evacuate the manometer or gauge, the connecting tubing, and the ICV to 10 mm Hg
absolute pressure. Close the flow-control and vacuum pump valves.
Begin auxiliary oxygen flow to the oxidation catalyst at a rate of 150
cc/min, then switch the four-way valve to the trap recovery position and the
sample recovery valve to collect position. The system should now be set up to
operate as indicated in Figure 25-10. After the manometer or pressure gauge begins
to register a slight positive pressure, open the flow control valve. Adjust the
flow-control valve to maintain atmospheric pressure in the system within 10
percent.
Now, remove the condensate trap from the dry ice, and allow it to warm to
ambient temperature while monitoring the NDIR response. If after 5 minutes, the
C02 concentration of the catalyst effluent is below 10,000 ppm, discontinue the
auxiliary oxygen flow to the oxidation catalyst. Begin heating the trap by placing
it in a furnace preheated to 200°C. Once heating has begun, carefully monitor the
NDIR response to ensure that the catalyst effluent concentration does not exceed
50,000 ppm. Whenever the C02 concentration exceeds 50,000 ppm, supply auxiliary
oxygen to the catalyst at the rate of 150 cc/min. Begin heating the tubing that
connected the heated sample box to the condensate trap only after the C02
concentration falls below 10,000 ppm. This tubing may be heated in the same oven
as the condensate trap or with an auxiliary heat source such as a heat gun.
Heating temperature must not exceed 200°C. If a heat gun is used, heat the tubing
slowly along its entire length from the upstream end to the downstream end, and
repeat the pattern for a total of three times. Continue the recovery until the C02
concentration drops to less than 10 ppm as determined by syringe injection as
described under the condensate trap C02 purge procedure, Section 4.3.2.
After the sample recovery is completed, use the carrier gas bypass valve to
pressurize the ICV to approximately 1060 mm Hg absolute pressure with zero air.
4.4 Analysis. Before putting the NMO analyzer into routine operation,
conduct an initial performance test. Start the analyzer, and perform all the
necessary functions in order to put the analyzer into proper working order; then
conduct the performance test according to the procedures established in Section
5.2. Once the performance test has been successfully completed and the C02 and NMO
calibration response factors have been determined, proceed with sample analysis as
follows:
4.4.1 Daily Operations and Calibration Checks. Before and immediately after
the analysis of each set of samples or on a daily basis {whichever occurs first),
conduct a calibration test according to the procedures established in Section 5.3-
If the criteria of the daily calibration test cannot be met, repeat the NMO
analyzer performance test (Section 5*2) before proceeding.
4.4.2 Operating Conditions. The carrier gas flow rate is 29.5 cc/min He and
2.2 cc/min 02 . The column oven is heated to 85°C. The order of elutiori for the
sample from the column is CO, CHA, C02, and NMO.
-------
Section No. 3.17.10
Date May 31, 1991
Page 10
4.4.3 Analysis of Recovered Condensate Sample. Purge the sample loop with
sample, and then inject the sample. Under the specified operating conditions, the
CO, in the sample will elute in approximately 100 seconds. As soon as the detector
response returns to baseline following the C02 peak, switch the carrier gas flow ,to
backflush, and raise the column oven temperature to 195°C as rapidly as possible.
A rate of 30°C/min has been shown to be adequate. Record the value obtained for
the condensible organic material (Ccm) measured as C02 and any measured NMO.
Return the column oven temperature to 85°C in preparation for the next analysis.
Analyze each sample in triplicate, and report the average Ccm.
4.4.4 Analysis of Sample Tank. Perform the analysis as described in Section
4.4.3, but record only the value measured for NMO (Ctm).
4.5 Audit Samples. Analyze a set of two audit samples concurrently with any
compliance samples and in exactly the same manner to evaluate the analyst's
technique and the instrument calibration. The same analysts, analytical reagents,
and analytical system shall be used for the compliance samples and the EPA audit
samples; if this condition is met, auditing of subsequent compliance analyses for
the same enforcement agency within 30 days is not required. An audit sample set
may not be used to validate different sets of compliance samples under the
jurisdiction of different enforcement agencies, unless prior arrangements are made
with both enforcement agencies.
Calculate the concentrations of the audit samples in ppm using the specified
sample volume in the audit instructions. (NOTE; Indication of acceptable results
may be obtained immediately by reporting the audit results in ppm and compliance
results in ppm by telephone to the responsible enforcement agency.) Include the
results of both audit samples, their identification numbers, and the analyst's name
with the results of the compliance determination samples in appropriate reports to
the EPA regional office or the appropriate enforcement agency during the 30-day
period.
The concentration of the audit samples obtained by the analyst shall agree
within 20 percent of the actual concentrations. Failure to meet the 20 percent
specification may require retests until the audit problems are resolved. However,
if the audit results do not affect the compliance or noncompliance status of the
affected facility, the Administrator may waive the reanalysis requirement, further
audits, or retests and accept the results of the compliance test. While steps are
being taken to resolve audit analysis problems, the Administrator may also choose
to use the data to determine the compliance or noncompliance of the affected
facility. .
5. Calibration and Operational Checks
Maintain a record of performance of each item.
5-1 Initial Performance Check of Condensate Recovery Apparatus. Perform these
tests before the system is first placed in operation, after any shutdown of 6
months or more, and after any major modification of the system, or at the specified
frequency.
5.1.1 Carrier Gas and Auxiliary 02 Blank Check. Analyze each new tank of
carrier gas or auxiliary 02 with the NMO analyzer to check for contamination.
Treat the gas cylinders as noncondensible gas samples, and analyze according to the
procedure in Section 4.4.3. Add together any measured CHA, CO, C02, or NMO. The
o
o
o
-------
Section No. 3.17.10
Date May 31, 1991
Page 11
total concentration must be less than 5 ppm.
5.1.2 Catalyst Efficiency Check. With a clean condensate trap installed in
the recovery system, replace the carrier gas cylinder with the high level methane
standard gas cylinder (Section 3 • **•!)• Set the four-port valve to the recovery
position, and attach an ICV to the recovery system. With the sample recovery valve
in vent position and the flow-control and ICV valves fully open, evacuate the
manometer or gauge, the connecting tubing, and the ICV to 10 mm Hg absolute
pressure. Close the flow-control and vacuum pump valves.
After the NDIR response has stabilized, switch the sample recovery valve from
vent to collect. When the manometer or pressure gauge begins to register a slight
positive pressure, open the flow-control valve. Keep the flow adjusted so that
atmospheric pressure is maintained in the system within 10 percent. Continue
collecting the sample in a normal manner until the ICV is filled to a nominal gauge
pressure of 300 mm Hg. Close the ICV valve, and remove the ICV from the system.
Place the sample recovery valve in the vent position, and return the recovery
system to its normal carrier gas and normal operating conditions. Analyze the ICV
for C02 using the NMO analyzer; the catalyst efficiency is acceptable if the C02
concentration is within 2 percent of the methane standard concentration.
5.1.3 System Performance Check. Construct a liquid sample injection unit
similar in design to the unit shown in Figure 25-7 • Insert this unit into the
condensate recovery and conditioning system in place of a condensate trap, and set
the carrier gas and auxiliary 02 flow rates to normal operating levels. Attach an
evacuated ICV to the system, and switch from system vent to collect. With the
carrier gas routed through the injection unit and the oxidation catalyst, inject a
liquid sample (See Sections 5.1-3«1 to 5-1-3-4) into the injection port. Operate
the trap recovery system as described in Section 4.'3«3« Measure the final ICV
pressure, and then analyze the vessel to determine the C02 concentration. For each
injection, calculate the percent recovery using the equation in Section 6.6.
The performance test is acceptable if the average percent recovery is 100 +_ 10
percent with a relative standard deviation (Section 6.9) of less than 5 percent for
each set of triplicate injections as follows:
5.1.3.1 50 ul Hexane.
5.1.3.2 10 ul Hexane.
5.1.3.3 50 ul Decane.
5.1.3.4 10 ul Decane.
5.2 Initial NMO Analyzer Performance Test. Perform these tests before the
system is first placed in operation, after any shutdown longer than 6 months, and
after any major modification of the system.
5.2.1 Oxidation Catalyst Efficiency Check. Turn off or bypass the NMO
analyzer reduction catalyst. Make triplicate injections of the high level methane
standard (Section 3«^-l)- The oxidation catalyst operation is acceptable if the
FID response is less than 1 percent of the injected methane concentration.
5.2.2 Reduction Catalyst Efficiency Check. With the oxidation catalyst
unheated or bypassed and the heated reduction catalyst bypassed, make triplicate
-------
Section No. 3.17.
Date May 31 , 1991
Page 12
injections of the high level methane standard (Section 3.4.1). Repeat this
procedure with both catalysts operative. The reduction catalyst operation is
acceptable if the response under both conditions agree within 5 percent.
5.2.3 Analyzer Linearity Check and NMO Calibration. While operating both the
oxidation and reduction catalysts, conduct a linearity check of the analyzer using
the propane standards specified in Section 3.4.2. Make triplicate injections of
each calibration gas, and then calculate the average response factor (area/ppm C)
for each gas, as well as the overall mean of the response factor values. The
instrument linearity is acceptable if the average response factor of each
calibration gas is within 2.5 percent of the overall mean value and if the relative
standard deviation (Section 6.9) for each set of triplicate injections is less than
2 percent. Record the overall mean of the propane response factor values as the
NMO calibration response factor (RFNMO).
Repeat the linearity check using the C02 standards specified in Section 3-4.3-
Make triplicate injections of each gas, and then calculate the average response
factor (area/ppm C) for each gas, as well as the overall mean of the response
factor values. Record the overall mean of the response factor values as the C02
calibration response factor (RFV,-. ) . Linearity is acceptable if the average
response factor of each calibratlbr? gas is within 2.5 percent of the overall mean
value and if the relative standard deviation for each set of triplicate injections
is less than 2 percent. The RFC_ must be within 10 percent of the RFNM0.
5.2.4 System Performance Check. Check the column separation and overalllJ
performance of the analyzer by making triplicate injections of the calibration
gases listed in Section 3-4.4. The analyzer performance is acceptable if the
measured NMO value for each gas (average of triplicate injections) is within 5
percent of the expected value.
5.3 NMO Analyzer Daily Calibration.
5.3-1 C02 Response Factor. Inject triplicate samples of the high level C02
calibration gas (Section 3-4.3). and calculate the average response factor. The
system operation is adequate if the calculated response factor is within 5 percent
of the RFpQ calculated during the initial performance test (Section 5.2.3). Use
the daily response factor (DFRCQ ) for analyzer calibration and the calculation of
measured C02 concentrations in tKe ICV samples.
5.3-2 NMO Response Factors. Inject triplicate. samples of the mixed propane
calibration cylinder (Section 3-4.4.1), and calculate the average NMO response
factor. The system operation is adequate if the calculated response factor is
within 5 percent of the RFNMO calculated during the initial performance test
(Section 5.2.4). Use the daily response factor (DRFNMO) for analyzer calibration
and calculation of NMO concentrations in the sample tanks.
5.4 Sample Tank and ICV Volume. The volume of the gas sampling tanks used
must be determined. Determine the tank and ICV volumes by weighing them empty and
then filled with deionized distilled water; weigh to the nearest 5 g, and record
the results. Alternatively, measure the volume of water used to fill them to
nearest 5 ml-
-------
Section No.
Date May 31,
Page 13
6. Calculations
3.17.10
1991
All equations are written using absolute pressure; absolute pressures are
determined by adding the measured barometric pressure to the measured gauge or
manometer pressure.
6.1 Nomenclature.
C = TGNMO concentration of the effluent, ppm C equivalent.
Cc = Calculated condensible organic (condensate trap) concentration
of the effluent, ppm C equivalent.
Ccm = Measured concentration {NMO analyzer) for the condensate trap
ICV, ppm C02.
Ct = Calculated noncondensible organic concentration (sample tank)
of the effluent, ppm C equivalent.
Ctm = Measured concentration (NMO analyzer) for the sample tank,
ppm NMO.
F = Sampling flow rate, cc/min.
L = Volume of liquid injected, ul.
M = Molecular weight of the liquid injected, g/g-mole.
mc = TGNMO mass concentration of the effluent, mg C/dsm3.
N = Carbon number of the liquid compound injected (N = 12 for
decane, N = 6 for hexane).
Pf = Final pressure of the intermediate collection vessel, mm Hg
absolute.
Pb = Barometric pressure, cm Hg.
Gas sample tank pressure before sampling, mm Hg absolute.
Gas sample tank pressure after sampling, but before
pressurizing, mm Hg absolute.
Final gas sample tank pressure after pressurizing, mm Hg
absolute.
Final temperature of intermediate collection vessel, °K.
Sample tank temperature before sampling, °K.
Sample tank temperature at completion of sampling, °K.
Ttf = Sample tank temperature after pressurizing, °K.
V = Sample tank volume, m3.
Vt = Sample train volume, cc.
Vv = Intermediate collection vessel volume, m3.
Vs = Gas volume sampled, dsm3.
n = Number of data points.
q = Total number of analyzer injections of intermediate collection
vessel during analysis (where k = injection number, 1 ... q).
r = Total number of analyzer injections of sample tank during
analysis (where j = injection number, 1 ... r).
xx = Individual measurements.
x = Mean value.
P = Density of liquid injected, g/cc.
6 = Leak check period, min.
AP = Allowable pressure change, cm Hg.
• ti
Pt
t f
ti
T
6.2 Allowable Pressure Change.
the allowable pressure change:
For the pretest leak check, calculate
-------
AP = 0.01
Section No. 3.17.10
Date May 31. 1991
Page 14
Eq. 25-1
6.3 Sample Volume. For each test run, calculate the gas volume sampled:
Eq. 25-2
v. = 0.3857 v
t _ * ti
o
6.4 Noncondensible Organics. For each sample tank, determine the
concentration of nonmethane organics (ppm C):
t r
t f
P P
rt rt
T T
1t At
i E ct
r 3=1
Eq. 25-3
6.5 Condensible Organics. For each condensate trap, determine the concentra
tion of organics (ppm C):
Eq. 25-4
-O
; Cc = 0.3857
V*,
1 V C
— /-J
cm,.
k=l
6.6 TGNMO. To determine the TGNMO concentration for each test run, use the
following equation:
c = ct + cc
Eq. 25-5
6.7 TGNMO Mass Concentration. To determine the TGNMO mass
concentration as carbon for each test run, use the following equation:
mc = 0.4993 C
Eq. 25-6
6.8 Percent Recovery. To calculate the percent recovery for the liquid
injections to the condensate recovery and conditioning system use the following
equation:
M vv rf c/cm
Percent recovery = 1.604 —
P T N
Eq. 25-7
O
-------
Section No. 3.17.10
Date May 31, 1991
Page 15
6.9 Relative Standard Deviation.
100 /r(x1 - :
- » n - 1
x
- x)2
RSD = "W Eq. 25-8
7. Bibliography
1. Salo, Albert E., Samuel Witz, and Robert D. MacPhee. Determination of
Solvent Vapor Concentrations by Total Combustion Analysis: A Comparison of
Infrared with Flame lonization Detectors. Paper No. 75~33-2. (Presented at the
68th Annual Meeting of the Air Pollution Control Association. Boston,
Massachusetts. June 15-20, 1975.) 14 p.
2. Salo, Albert E., William L. Oaks, and Robert D. MacPhee. Measuring the
Organic Carbon Content of Source Emissions for Air Pollution Control. Paper No.
74-190. (Presented at the 67th Annual Meeting of the Air Pollution Control
Association. Denver, Colorado. June 9-13, 1974.) 25 p.
-------
Section No.
Date May 31,
Page 16
3.17.10
1991
o
REGULATING
VALVE
MANOMETER
DUAL RANGE
ROTAMETER
TEMPERATURE
CONTROLLER
VACUUM PUMP
THERMOCOUPLES
PURGE VALVE
THERMOCOUPLE
ROTAMETER | |
STACK
WALL
SAMPLE
TANK
VALVE
SAMPLE
VALVE
STAINLESS STEEL
FILTER HOLDER
HEATED BOX
CONDENSATE
TRAP
STAINLESS
STEEL PROBE
O
SAMPLE
TANK
Figure 25-1. Sampling train.
O
-------
Section No. 3.17.10
Date May 31, 1991
Page 17
VACUUM PUMP
CONNECTOR
SAMPLE
25.4
1.0
F1BERFAX
INSULATION
DIMENSIONS: HE1
in
3.175
0.125
CONDENSATE
TRAP PROBE
BULKHEAD
CONNECTOR
J
PROBE
NNECTOR
PROBE LINE
THERMOCOUPLE
TO TEMPERATURE
CONTROLLER
a a
FILTER HEAT CONDENSATE
TEMPERATURE TRAP PROBE
CONTROLLER CONNECTOR
THERMOCOUPLE THERMOCOUPLE
Figure 25-2. Out-of-stack filter box.
. \ t_-
-------
Section No.
Date May 31,
Page 18
3.17.10
1991
o
DIMENSIONS:
nrn
in
0.375 0.035
316SS TUBING
31 ess NUT
WALL
O
COARSE QUARTZ
WOOL PACKING
2.25
Figure 25-3. Condensate trap.
O
-------
eftf
Section No. 3-17.10
Date May 31. 1991
Page 19
FLOW METERS
HEATTRACE(100°C)
SAMPLE
RECOVERY
VALVE
FLOW
CONTROL
VALVE
SYRINGE PORT
VACUUM PUMP
Figure 25-4. Condensate recovery system.
-------
Section No. 3.17.10
Date May 31, 1991 S~\
Page 20 (J
CARRIER GAS
CALIBRATION STANDARDS —1>
SAMPLE TANK
i
SAMPLE
INJECTION
LOOP
INTERMEDIATE COLLECTION
<3— VESSEL (CONDmONED
TRAP SAMPLE)
SEPARATION
COLUMN
BACKFLUSH
CO.CH4lC02
NONMETHANE
ORGANICS
OXIDATION
CATALYST
REDUCTION
CATALYST
FLAME
IONIZATION
DETECTOR
HYDROGEN
COMBUSTION
AR
DATA RECORDER
o
Figure 25-5. Simplified schematic of nonmelhane organic (NMO) analyzer.
O
, •)
-------
Section No. 3.17-10
Date May 31, 1991
Page 21
COLUMN OVEN
rOQOOOOOOOOOOOQQQOOOQQQ
REDUCTION
CATALYST
H2 AIR
Figure 25-6. Nonmethane organic (NMO) analyzer.
-------
CONNECTING T
INJECTION
SEPTUM
Section No.
Date May 31,
Page 22
CONNECTING ELBOW
3.17.10
1991
FROM
CARRIER
DIMENSIONS:
o
. TO
CATALYST
O
31 eSS TUBING
Figure 25-7. Liquid sample injection unit.
O
8/89
-------
Section No. 3-17-10
Date May 31, 1991
Page 23
VOLATILE ORGANIC CARBON
FACILITY
LOCATION
DATf
TAWXNUMBfn
SAMPLE IOCAT(OM
OPERATOR
OHM UllUf CO
TBAFWUMBEH «AU»1 f 10 MUMBFH
"V : TAWXVACUUU.
.: mm Hf cm Hf
PRETEST (MANOMETER)
POST TEST (MANOMETER)
ICAUCfJ
(RAUCri
• BAROMETRIC
PRESSURE.
mm Nf
AMBIENT
TIUPtRATURE,
•c
LEAK RATE
cm H| / 10 min
PRETEST.
TIME
CLOCK/SAMPLE
VACUUM
cm H|
flOWMETIR SITTING
COMMENTS
Figure 25-8. Example field data form.
-------
Section No. 3.17.10
Date May 31, 1991
Page 2H
o
FLOW METERS
AIR
HEAT TRACE (100'C)
n
H2OTRAP
4-PORT VALVE
CONDENSATEi
TRAP
FLOW
CONTROL
VALVE
C_DRYICE_^
VENT
SAMPLE
RECOVERY
VALVE
NDIR
ANALYZER
LJ^LJ u-**1—'n' ^ ^<
OXIDATION
REACTOR
SYRINGE PORT
C7
VACUUM PUMP
SAMPLE
TANK
VALVE
SAMPLE
TANK
O
Figure 25-9. Condensate recovery system, CO2 purge.
O
-------
Section No. 3.17-10
Date May 31, 1991
Page 25
FLOW METERS
HEATTRACE(100°C)
SAMPLE
RECOVERY
VALVE
FLOW
CONTROL
VALVE
SYRINGE PORT
VACUUM PUMP
Figure 25-10. Condensate recovery system, collection of trap organics.
-------
o
o
o
-------
Section No. 3.17.11
Date May 31. 1991
Page 1
11.0 REFERENCES
1. Evaluation of Trap Recovery Design, EMB Project Number 82SFS-1.
2. Preparation of Method 25 Sampling Equipment and Determination of Linit of
Detection of Quantification, EMB Project No. 82SFS-1.
3. Evaluation of Method 25 Condensate Trap Packing Material, EMB Project Number
82SFS-1.
4. Oxidation Catalyst Screening and Evaluation Study, ESED Project Number 82SFS1-4-
2.
5. Quality Control Procedures Evaluation, ESED Project Number 82SFS1-4-3.
6. Condensate Trap Development and Evaluation, ESED Project Number 82SFS1-4-4.
•\
7. Trap Recovery Procedures Evaluation. ESED Project Number 82SFS1-4-5.
8. Evaluation of Particulate Filters, ESED Project Number 82SFS1-5-2.
9. "Procedure for NBS-Traceable Certification of Compressed Gas Working Standards
Used for Calibration and Audit of Continuous Source Emission Monitors (Revised
Traceability Protocol No. 1)," June 1987, Section 3.0.4 of the Quality Assurance
Handbook for Air Pollution Measurement Systems, Volume III, Stationary Source
Specific Methods, EPA-600/4-77-027b, August 1977. U. S. Environmental Protection
Agency, Office of Research and Development Publications, 26 West St. Clair St.,
Cincinnati, OH ^5268.
10. "A Procedure for Establishing Traceability of Gas Mixtures to Certain National
Institute for Standards and Technology Standard Reference Materials," Joint
Publication by NIST and EPA, EPA-600/7-81-010, Available from U. S.
Environmental Protection Agency, Quality Assurance Division (MD-77). Research
Triangle Park, North Carolina 27711.
11. R. S. Wright, C. V. Wall, C. E. Decker, and D. J. von Lehmden, "Accuracy
Assessment of EPA Protocol Gases in 1988," Journal of the Air and Waste
Management Association. 29_ (9): 1225-1227, September 1989.
-------
Section No 3.19
Date September 3, 1952
Page 1
SUMMARY
Method 101A, for determining particulate and gaseous mercury (Hg) emissions
from stationary sources, is similar to Method 101. In 101A, however, acidic potassium
permanganate (KMnO<) solution is used for sample collection instead of acidic iodine
monochloride. This method applies to determining particulate and gaseous mercury (Kg)
emissions from sewage sludge incinerators and other sources as specified in the
regulations. Particulate and gaseous Hg emissions are withdrawn isokinetically from
the source and collected in an acidic KMnO4 solution. The collected Hg (in mercuric
form) is reduced to elemental Hg, which is then aerated from the solution into an
optical cell and measured by atomic absorption spectrophotometry (AAS).
After initial dilution, the range of this method is 20 to 800 ng Hg/mL. The
upper limit can be extended by further dilution of the sample. The sensitivity of the
method depends on the recorder/ spectrophotometer combination selected. The collection
efficiency of the sampling method can be affected by excessive oxidizable matter in the
stack-gas that prematurely depletes the KMn04 solution.
The method descriptions given are based on the method1'2'3 promulgated October
15, 1980, and on corrections and additions published on September 12, 1984, and
September 23, 1988 (Section 3.19.10).
-------
o
o
o
-------
Section No 3.19
Date September 3, 1992
Page 1
METHOD HIGHLIGHTS
Section 3.19 describes the procedures and specifications for determining
particulate and gaseous mercury emissions from sewage sludge incinerators and other
stationary sources as specified in the regulations. New procedures were added to
Method 101A3 on the basis of EPA-conducted development and evaluation of mercury
sampling and analysis. The major changes for Method 101A are:
1. The impinger KMnO4 absorbing solution and the 8 N hydrochloric acid
(HCl) rinse are no longer combined in the field during sample recovery.
2. The impinger KMn04 absorbing solution must be filtered.
3. The filtrate must be analyzed within 24 h of filtration.
4. The residue on the filter from the filtration step must be digested .with
8 N HCl.
5. The HCl digestate and the final field sample recovery rinse qf HCl are
combined and analyzed separately from the KMnO4 filtrate. :;
1. Procurement of' Apparatus and Supplies ; ,, . •,,. -,.....
Section 3.19.1 (Procurement of Apparatus and Supplies) gives specifications,
criteria, and design features for the equipment find materials required for'Method 101A.
This section can be used as a guide for procuring and initially checking equipment and
supplies. The activity matrix (Table 1.1) at the end of the section is a summary of
the details given in the text and can be used as a quick reference.
2. Pretest Preparations
Section 3.19.2 (Calibration of Apparatus) describes the required calibration
procedures and considerations for the Method 101A sampling equipment. Required
accuracies for each component also are included. A pretest checklist (Figure 3.1 in
Subsection 3.19.3) or a similar form should be used to summarize the calibration and
other pertinent pretest data. The calibration section may be removed along with the
corresponding sections for the other methods and made into a separate quality assurance
reference manual for personnel involved in calibration activities. : -.
Section 3.19.3 (Presampling Operations) provides testers with a guide for
preparing equipment and supplies for field tests. A pretest preparation form can be
used as an equipment checkout and packing list. Because of the potential for high
blank levels, special attention must be paid to preparing the sampling equipment.
Also, testers must ensure that any required audit samples are obtained for the test by
the responsible regulatory agency.
Activity matrices for calibrating the equipment and the presampling operations
(Tables 2.1 and 3.1) summarize the activities.
3. On-Site Measurements
Section 3.19.4 (On-Site Measurements) contains step-by-step procedures for
sample collection, sample recovery, and sample preparation for transport. The on-site
checklist (Figure 4.3, Section 3.19.4) provides testers with a quick method for
checking the on-site requirements. Table 4.1 provides an activity matrix for all
on-site activities. c -
-------
Section No 3.19
Date September 3, 1992
Page 2
o
4. Posttest Operations
Section 3.19.5 (Posttest Operations) presents the posttest equipment
procedures and a step-by-step analytical procedure for determination of mercury, .
Posttest calibrations are required for the sampling equipment. The posttest operations,
form (Figure 5.9, Section 3.19.5) provides some key parameters that testers and
laboratory personnel must check. The step-by-step sample preparation ,and analytical
procedure descriptions can be removed and made into a separate quality assurance
analytical1reference manual for laboratory personnel.
Section 3.19.6 (Calculations) provides testers with the required equations,
nomenclature, and significant digits. A calculator or computer should be used, if
available, to reduce the chances of error.
Section 3.19.7 (Maintenance) provides testers with a guide for a maintenance
program. This program is not required, but it should reduce equipment malfunctions.
Activity matrices (Tables 5.1, 6.1, and 7.1) summarize all postsampling, calculation,
and maintenance activities.
5. Auditing Procedures
Section 3.19.8 (Auditing Procedure) provides a description of necessary
activities for conducting performance and system audits. The data-processing.*—^
procedures and a checklist for a systems audit also are included in this section/ j
Table 8.1 is an activity matrix for conducting the performance and system audits. \^S
Section 3.19.9 (Recommended Standards for Establishing Traceabi-lity) provides
the primary standard to which the analytical data should be traceable.
6. References
Section 3.19.10 contains the promulgated Method 101A; Section 3.19.11 contains
the references cited throughout the text.
O
-------
Section No. 3.19.1
Date September J, 1992
Page 1
1.0 PROCUREMENT OP APPARATUS AND SUPPLIES
Before Method 101A can yield results, it must be employed accurately.
Consequently, all users are advised to read this document and to adopt its procedures.
Alternative procedures should be employed only if they are outlined herein.
This section describes equipment specif ications/'briteria, and design features
for the sampling train used for Method 101A. It is intended to help users with
equipment.selection. A schematic of the sampling train is shown in Figure 1.1 as an
aid in the discussion that follows.
This section also describes procedures and limits, where applicable, for
acceptance checks. Calibration data generated by the acceptance checks should be
recorded in the calibration log book.
When procuring equipment and supplies, users should record the descriptive
title of the equipment, identification number (if applicable), and the results cf
acceptance checks in a procurement log. .._ ,
The following procedures and descriptions are provided only as guidance and
may not be required for the initial ordering and check-out of the equipment ar.d
supplies. Testers should note, however, that many of these procedures are required at
a later step in the sampling and analytical procedures. Instituting these or similar
procedures as routine practices for checking new equipment and supplies, therefore,
will prevent later problems and/or delays in test programs. At the end of this
section, Table 1.1 provides a summary of quality assurance activities for procurement
and acceptance of apparatus and supplies.
1.1 Sampling :
The sampling train shown in Figure 1.1 is similar to the Method 5 train
(Method 5 refers to 40 CFR Part 60). The Method 101A sampling train consists of the
following components:
1.1.1 Nozzle—The nozzle shall be made of nickel, nickel-plated stainless-steel,
quartz, or borosilicate glass. The tapered angle should be <30°, with taper on the
outside to preserve a constant inside diameter (ID). • . . :.
A range of nozzle ID's-for example, 0.3,2 to 1.27 cm (0.125 to 0.5 in.)-in
increments of 0.16 cm (0.0625 in.) should be available for isokinetic sampling. Larger
nozzle sizes may be required if very low flows are encountered.
Upon receipt of the nozzle(s) from the manufacturer, users should inspect it
for roundness, for the proper material, and for damage to the tapered edge (nicks,
dents, and burrs). Check the diameter with a micrometer; calibration procedures are
described in Section 3.18.2. A slight variation from exact sizes is normal. Engrave
each nozzle with an identification number for inventory and calibration purposes. See
Section 3.18.3 for proper cleaning procedures.
-------
HEATED AREA
TEMPERATURE
SENSOR
PROBE
THERMOMETER
\ FILTER HOLDER
T (OPTIONAL) THERMOMETER
CHECK
VALVE
TYPE S
PITOT TUBE
PITOT MANOMETER
IMPINGERS
THERMOMETERS
TEMPERATURE SENSOR
BY-PASS VALVE
f
PROBE
CJ>
1
PITOT TUBE
ICE BATH
MAIN
VALVE
/
VACUUM
GAUGE
»o o w
DI p> ro
(Q rr o
(0 0> rr
KJ tn o
VACUUM »z
(0 o
DRY G
METER
AIR TIGHT
PUMP
LINE
U)
.
to
o
Figure 1.1. Schematic of Met]wd 101A sampling train.
10
U3
to
O
-------
Section No. 3.19.1
Date September 3, 1992
Page 3
.1.2 Pi tot Tube-The pitot tube, preferably of Type S design, should meet the
requirements of Method 2, Section 3.1.2 of this Handbook. The pitot tube is attached
to the probe as shown in Figure 1.1. The proper pitot-tube/sampling-nozzle
configuration to prevent aerodynamic interference is shown in figures 2.6 and 2.7 of
Method 2, Section 3.1.2 of this Handbook.
The pitot tube should be inspected visually for both vertical and horizontal
tip alignments. If the tube is purchased as an integral part of a probe assembly,
check the dimensional clearances using figures 2.6 and 2.7 (of Method 2, Section
3.1.2). Repair or return any pitot tube that does not meet specifications. The
calibration procedure for a pitot tube is covered in Section 3.4.2 of this Handbook.
1.1.3 Differential Pressure AP-The differential pressure gauge should be an inclined
manometer or the equivalent, as specified in Method 2, Section 3.1.2 of this Handbook.
Two gauges are required. One is used to monitor the stack velocity pressure, whereas
the other is used to measure the orifice pressure differential.
Initially, check the gauge against a gauge-oil manometer at a minimum of three
points: 0.64 mm (0.025 in.); 12.7 mm (0.5 in.); and 25.4 mm (1.0 in.) H20. The gauge
should read within 5% of the gauge-oil manometer at each test point. Repair or return
to the supplier any gauge that does not meet these requirements.
1.1.4 Probe Liner—The probe liner is made of borosilicate or quartz glass tubing.
(Note: Do not use metal probe liners.) If a filter is used ahead of the impingers,
testers must use the probe heating system to minimize the condensation of gaseous Hg.
A heating system is required that will maintain an exit gas temperature of 120 ± 14 °C
k(248 ± 25 °F) during sampling. Other temperatures may be specified by a subpart of the
egulations and must be approved by the Administrator for a particular application.
ecause the actual probe outlet temperature is not usually monitored during sampling,
probes constructed in accordance to APTD-0581 and calibrated according to procedures
in APTD-0576 will be acceptable.
Either borosilicate or quartz glass liners may be used for stack temperatures
up to about 480 °C (900 °F), but quartz glass liners must be used from 480 to 900 °C
(900 to 1650 °F). Either type of liner may be used at higher temperatures for short
periods, with Administrator approval. However, the absolute upper limits—the softening
temperatures of 820 °C (1508 °F) and 1500 °C (2732 °F)-for borosilicate and quartz,
respectively, must be observed.
Upon receiving a new probe, users should check it visually to see whether it
is the length and composition ordered. The probe also should be checked visually for
breaks or cracks, and it should be checked for leaks on a sampling train (Figure 1.1).
Leak checks should include a proper nozzle-to-probe connection with a Viton O-ring,
Teflon® ferrules, or asbestos string.
The probe heating system should be checked as follows:
1. With a nozzle attached, connect the probe outlet to the inlet of the
metering system.
2. Connect the probe heater to an outlet and turn it on for 2 or 3 min.
The probe should become warm to the touch.
3. Start the pump and adjust the needle valve until it indicates a flow
rate of about 0.02 mVmin (0.75 ftVmin) .
4. Be sure the probe remains warm to the touch; the heater should be
capable of maintaining an exit air temperature of 100 °C (212 ^F)
minimum. Failure indicates that the probe should be repaired, returned
to the supplier, or rejected.
-------
Section No. 3.19.1
Date September 3, 1992
Page 4
1.1.5 Filter Holder (Optional;-The filter holders should be made of borosilicate
glass with a rigid, stainless-steel wire-screen filter support. Do not use glass frit
supports. A silicone rubber or Teflon gasket is essential to provide a positive seal
against leakage from outside or around the filter. Upon receipt, assemble the filter
holder with a filter and conduct a leak check. There should be no leak at a vacuum of
15 in. of Hg.
1.1.6 Impingers-Four Greenburg-Smith impingers must be connected in series with
leak-free, ground glass fittings or any similar leak-free, noncontaminating fittings.
For the first, third, and fourth impingers, testers may use impingers that are modified
by replacing the tip with a 13-mm ID (0.5 in.) glass tube extending to 13 mm (0.5 in.)
from the bottom of the flask. The connecting fittings should form leak-free,
vacuum-tight seals. See Section 3.19.3 for proper cleaning procedures.
Upon receipt of a standard Greenburg-Smith impinger, users should fill the
inner tube with water. If the water does not drain through the orifice in 6 to 8 s or
less, the impinger tip should be replaced or enlarged to prevent an excessive pressure
drop in the sampling system. Each impinger should be checked visually for damage:
.breaks, cracks, or manufacturing flaws, such as poorly shaped connections.
1.1.7 Acid Trap-The acid trap should be a Min~e Safety Appliances™ airline filter,
catalog number 81857, with acid absorbing cartridge and suitable connections, or the
equivalent. Upon receipt, check the part number to ensure the part is correct.
1.1.8 Filter Heating System-Any heating system may be used that is capable of
maintaining the filter holder at 120 ± 14 °C (248 ± 25 °F) during sampling. Other
temperatures may be specified by a subpart- of the regulations or approved by the
Administrator for a particular application. A gauge capable of measuring temperatures
to within 3 °C (5.4 °F) should be provided to monitor the temperature around the filter
during sampling.
Before sampling, the heating system and the temperature monitoring device
should be checked. For convenience, the heating element should be easily replaceable
should a malfunction occur during sampling.
1.1.9 Metering System-The metering system should consist of a vacuum gauge, a vacuum
pump, thermometers capable of measuring ±3 °C (5.4 °F) of true value in the range of
0 to 90 °C (32 to 194 °F) , a dry-gas meter with 2% accuracy at the required sampling
rate, and related equipment as shown in Figure 1.1. Other systems capable of
maintaining metering rates within 10% of the isokinetic sampling rate and of
determining sample volumes to within 2% of the isokinetic rate may be used if approved
by the Administrator. Sampling trains with metering systems designed for sampling
rates higher than those described in APTD-0581 and APTD-0576 may be used if the above
specifications can be met. When the metering system is used with a pitot tube, it
should permit verification of an isokinetic sampling rate through the use of a
nomograph or by calculation.
Upon receipt or after construction of the system, users should perform both
positive and negative pressure leak checks before beginning the system calibration
procedure described in Subsection 2.1 of Section 3.19.2. Any leakage requires repair
or replacement of the malfunctioning item.
i
1.1.10 Barometer—A mercury, aneroid, or other barometer capable of measuring
atmospheric pressure to within ± 2.5 mm (0.1 in.) Hg is required.
A preliminary check of a new barometer should be made against a
mercury-in-glass barometer or the equivalent. In lieu of a barometer check, the
o
o
o
-------
Section No. 3.19.1
Date September 3, 1992
Page 5
absolute barometric pressure may be obtained from .a nearby weather service station and
adjusted for elevation difference between the station and the sampling point. Either
subtract 2.5 mm Hg/30 m (0.1 in. Hg/100 ft) from the station value for an elevation
increase, or add the same for an elevation decrease. If the barometer cannot be
adjusted to within 2.5 mm {0.1 in.) Hg of the reference barometric pressure, it should
be returned to the manufacturer or rejected.
1.1.11 Gas Density Determination Equipment—& temperature sensor and a pressure gauge
are required as described in Method 2 (Section 3.1.2 of this Handbook). Additionally,
a gas analyzer as described by Method 3 may be required. The temperature sensor should
be permanently attached to either the probe or the pitot tube. In either case, it is
recommended that a fixed configuration (Figure 1.1) be maintained. The sensor also may
be attached just prior to field use, as described in Section 3.19.2.
1.2 Sample Recovery
1.2.1 Glass Sample Bottles Sample bottles should be 1000- and 100-mL without leaks
and with Teflon-lined caps. Upon receipt, check visually for cracks in the glass.
Ensure that the cap liners are Teflon, because other material can result in sample
contamination and reaction with the KMnO«. Beca~use of the potential reaction of the
04 with the acid, there may be pressure buildup in the sample storage bottles.
enting is highly recommended. A No. 70-72 hole drilled in the container cap and
Teflon liner has been found to allow adequate venting without loss of sample.
1.2.2 Graduated Cylinder—A 250-mL cylinder is required.
1.2.3 Funnel and Rubber Policeman—These items are used to aid in transferring silica
gel to containers; they are not necessary if silica gel is weighed in the field.
1.2.4 Funnel—A glass funnel is needed to aid in sample recovery.
1.3 Sample Preparation and Hg Analysis
1.3.1 Volumetric Pipets-Class A 1-, 2-, 3-, 4-, 5-, 10-, and 20-mL pipets are
required.
1.3.2 Graduated Cylinder—A 25-mL cylinder is required.
1.3.3 Steam Bath-Refers to 40 CFR, Part 60, Appendix B, Method 101A.
1.3.4 Atomic Absorption Spectrophotometer—Any atomic absorption unit is suitable,
provided it has an open sample presentation area in which to mount the optical cell.
Follow the instrument settings recommended by the manufacturer. Instruments designed
specifically for measuring mercury using the cold-vapor technique are commercially
available and may be substituted for the atomic absorption spectrophotometer.
1.3.5 Optical Cell—The optical cell should be of cylindrical shape, with quartz end
windows and having the dimensions shown in Figure 1.2. Wind the cell with
approximately 2 m of 24-gauge nichrome heating wire, and wrap with fiberglass
insulation tape or the equivalent; do not let the wires touch each other. A heat lamp
mounted above the cell or a moisture trap installed upstream of the cell may be used
as alternatives. Upon receipt, check the dimensions and the capability of the heating
system.
-------
Section No. 3.19.1
Date September 3, 195
Page 6
o
1.3.6 Aeration Cell—The aeration cell should be constructed according to the
specifications in Figure 1.3. Do not use a glass frit as a substitute for the blowr.
glass bubbler tip shown in Figure 1.3.
1.3.7 Recorder--The recorder must be matched to the output of the spectrophotometer
described above. As an alternative, an integrator may be used to determine peak area
or peak height.
1.3.8 Variable Transformer-This item is needed to vary the voltage on the optical
cell from 0 to 40 volts.
1.3.9 Hood-A hood i's-) required for venting the optical cell exhaust.
1.3.10 Flow Metering Valve-Refers to 40 CFR,.. Part 60, Appendix B, Method 101A.
1.3.11 Flow Meter—A rotameter, or equivalent, is required that is capable cf
measuring a gas flow of 1.5 L/min. Upon receipt, calibrate the flow meter at a flow
rate of 1.5 L/min with a bubble meter or wet-test meter.
1.3.12 Aeration Gas Cylinder-The cylinder mustncontain nitrogen or dry, Hg-free air
and must be equipped with a single-stage regulator.
O
-o
-------
Section No. 3.19.1
Date September 3, 1992
Page 7
18/9 FEMALE BALL SOCKET
LENGTH NECESSARY TO FIT SOLUTION CELL
TO SPECTROPHOTOMETER
(END VIEW)
TO VARIABLE TRANSFORMER
VENT TO HOOD
4
9-mm 00
9-mm OD ,•£*& 2JS em
3.81 em DIAMETER
QUARTZ WINDOWS
AT EACH END
(FRONT VIEW)
NOTES:
CELL WOUND WITH 24-GAUGE NICHROME WIRE
TOLERANCES ± 5 PERCENT
Figure 1.2. Optical cell.
-------
Section No. 3.19.1
Date September 3, 1992
Page 8
o
FROM TANK
1 /"T"V 1B/B MALE BALL JOINT
T(r—7T 4-mmDORE TEFLON STOPCOCK
I IT innnnien _/ "^^
10/72 GROUND
10/22 GROUND
CLASS JOINT
WITH STOPPER
TO
OPTICAL CELL
18/8 MALE OALL JOINT
ALL DIMENSIONS IN em
UNLESS OTHERWISE NOTED
BLOWN GLASS BUMUIR COTTU POBT\ON
AH-ROX. &0 by 1.0 em 4.0-«m OD by 3.C-cm ID
O
o
Figure 1.3. Aeration cell.
-------
Section No. 3.19 .1
Date September 3, 1992
Page 9
1.3.13 TuJbing-The tubing is required for connections. Use glass tubing (ungreased
ball and socket connections are recommended) for all connections between the solution
cell and the optical cell; do not use Tygon tubing, other types of flexible tubing, or
metal tubing as substitutes. Testers may use Teflon, steel, or copper tubing between
the nitrogen tank and the flow meter valve (Section 5.3.7), and Tygon, gum, or rubber
tubing between the flow meter valve and the aeration cell.
1.3.14 Flow Rate Calibration Equipment—This equipment consists of a bubble flow meter
or a wet-test meter for measuring a gas flow rate of 1.5 ± 0.1 L/min.
1.3.15 Volumetric Flasks—These flasks must be Class A, with pennyhead standard taper
stoppers; the required sizes are 100-, 250-, 500-, and 1000-mL.
1.3.16 Volumetric Pipets—These pipets must be Class A; the required sizes are 1-, 2-,
3-, 4-, and 5-mL.
1.3.17 Graduated Cylinder- A 50-mL cylinder is required.
1.3.18 Magnetic Starrer— A general purpose laboratory-type stirrer is required.
-i
1.3.19 .Magnetic Stirring Bar— A Teflon-coated stirring bar is required.
1.3.20 Trip Balance— A trip balance capable of weighing to ± 0.5 g is required. Upon
receipt, check balance with standard weights.
1.3.21 Analytical Balance-An analytical balance capable of weighing up to ± 0.5 me
is required. Upon receipt, check balance with standard weights.
1.4 Alternative Analytical Apparatus
If any alternative analytical apparatus is to be used, it must pass the
performance criteria described in Section 3.19.5.5. Alternative Hg cold-vapor
analytical systems are available commercially from most atomic absorption manufacturers
and employ automated flow-injection techniques. Such systems automatically inject
sample solutions into continuous reagent streams containing the reducing reagent.
Mercury is usually measured as a solution concentration (e.g., mg Hg/L). An example
of a typical cold-vapor AA instrument using flow injection is shown in Figure 1.4.
Such systems are allowable as long as they meet the following criteria:
1.4.1 Calibration Curve Linearity—The system must generate a linear calibration
curve, and two consecutive samples of the same aliquot size and concentration must
agree within 3% of their average.
-------
oooo
oooo
oooo
oooo
Autosampler
Multichannel Pump
Inert Gas
Dilute HC1
Sample
H2S04
Reducing Reagent
Mixing Colls
Spectrophotometer w/ Optical Cell
Gas/Liquid Separator
To Vent
To Waste
Figure 1.4. Typical cold Vapor AA instrumentation using flow injection.
O O
•o O 0
o-O P
U) U)
vo
to
o
-------
Section No. 3.19.1
'- • Date September 3, 1992
'•' Page 11
1.4.2 Spike flecovery-The system must allow for recovery of a minimum of 95% of the
spike when an aliquot of a source sample is spiked with a known concentration of Hg
(II) compound.
1.5 Reagents
1.5.1 Sampling and Sample .Recovery-Use ACS reagent-grade chemicals or the
equivalent, unless otherwise specified. The following reagents are used in sampling
and recovery: -----
Water—Deionized distilled, , meeting ^ASTM specifications for Type I Reagent
Water—ASTM Test Method D 1193-77. If high concentrations'of organic matter are not
expected to be present, users may eliminate the KMnO< test for oxidizable organic
matter. Use this water in all dilutions and solution preparations.
Nitric Acid (HNO3), 50% (v/v)-Mix-equal volumes of concentrated HNO3 and water,
being careful to add the acid to the water slowly.
Silica Gel—Indicating type, 6- to 16-mesh. If previously used, dry at 175 °C
(350 °F) for 2 h. Testers may use new silica gel .as received.
Filter (Optional)—Glass fiber filter, without organic binder, exhibiting at
least 99.95% efficiency on 0.3-Hm dioctyl phthalate smoke particles. Testers may use
the filter in cases where the gas stream contains large quantities of particulate
matter, but they should analyze blank filters for Hg content.
Sulfuric Acid (H2SO4), 10% (v/v;-Slowly add 100 mL of concentrated H2SO< to 500
mL of water and mix cautiously.
Absorbing Solution, 4% KMnO, (w/v)— Prepare fresh daily. Dissolve 40 g of KMn04
in sufficient 10% H2SO4 to make 1 L. Prepare and store in glass bottles to prevent
degradation.
Caution: To prevent autocatalytic decomposition of the permanganate solution, filter
it through Whatman" 541 filter paper. In addition, owing to the reaction of the KKnO,
with the acid, there may be pressure buildup in the sample storage bottle. These
bottles should not be filled to capacity and should be vented, both to relieve excess
pressure and to prevent explosion of the container: A No. 70-72 hole drilled in the
container cap and Teflon liner is recommended.
Hydrochloric Acid-Trace metals grade HC1 is recommended. If other grades sre
used, the Hg level must be less than 3 ng/mL Hg. Upon receipt, check manufacturer's
guarantee or analyze the acid for background contamination.
Hydrochloric Acid, 8 AHDilute 67 mL of concentrated HC1 to 100 mL with water
(slowly add the HC1 to the water).
-------
Section No. 3.19.1
Date September 3, 1992
Page 12
1.5.2 Analysis—The reagents needed for analysis are listed below:
Tin (II) Solution—Prepare fresh daily and keep sealed when not being used.
Completely dissolve 20 g of tin (II) chloride [or 25 g of tin (II) sulfate] crystals
(Baker™ Analyzed reagent grade or any other brand that will give a clear solution) in
25 mL of concentrated HC1. Dilute to 250 mL with water. Do not substitute HN03, H2SO4,
or other strong acids for the HC1.
Sodium Chloride-tfydroxyl&mine Solution-Dissolve 12 g of sodium chloride and
12 g of hydroxylamine sulfate (or 12 g of hydroxylamine hydrochloride) in water and
dilute to 100 mL.
Hydrochloric Acid, 8 AMDilute 67 mL of concentrated HC1 to 100 mL with water
(slowly add the HCl to the water).
Nitric Acid, 15% (v/v)— Dilute 15 mL of concentrated HN03 to 100 mL with water.
Mercury Stock Solution, 1 mg Hg/mL-Prepare'and store all Hg standard solutions
in borosilicate glass containers. Completely dissolve 0.1354 g of Hg (II) chloride in
75 mL of water. Add 10 mL of concentrated HN03-~and adjust the volume to exactly 100
mL with water. Mix thoroughly. This solution is stable for at least 1 month.
Intermediate Hg Standard Solution, 10 ug/mL—Prepare fresh weekly. Pipet 5.0
mL of the Hg stock solution (Section 6.2.5) into a 500-mL volumetric flask, and add 20
mL of 15% HNO, solution. Adjust the volume to exactly 500 mL with water. Thoroughly
mix the solution.
Working Hg Standard Solution, 200 ng Jfg/mL—Prepare fresh daily. Pipet 5.0 mL
from the Intermediate Hg Standard Solution (Section 6.2.6) into a 250-mL volumetric
flask. Add 5 mL of 4% KMnO< absorbing solution and 5 mL of 15% HNOj. Adjust the volume
to exactly 250 mL with water. Mix thoroughly.
Potassium Permanganate, 5* (w/v)—Dissolve 5 g of KMn04 in water and dilute to
100 mL.
Filter-Use a Whatman 40, or equivalent.
o
o
o
-------
Section No. 3.19.1
Date September 3, 1992
Page 13
TABLE 1.1 ACTIVITY MATRIX FOR PROCUREMENT OF APPARATUS AND SUPPLIES
Apparatus
|Acceptance limits
I Frequency & method
I of measurement
I Action if
requirements
are not met
Sampling
Probe liner
V ,
I
(specified material
|of construction;
|equipped with
(heating system
|capable of
(maintaining 120 °C
|± 14°C (248 °C ±
125 °F) at the exit
1
(Visually check and
(run the heating
(system
I
Repair,
return to
supplier, or
reject
Probe
nozzle
(Nickel, nickel-
|plated stainless-
|steel, quartz, or
(borosilicate
(glass, tapered <
(30°; difference in
(measured diameter
|< 0.1 mm (0.004
I in.); no nicks,
(dents, or
(corrosion
I(Subsec. 1.1.2)
I ____
(Visually check
(before each test;
(use a micrometer to
(measure ID before
|field use; after
|each repair
I
I
Reshape and
sharpen,
return to"
the supplier,
or reject
Pi tot tube
(Type S (Sec.
3.1.2); attached
to probe with im-
pact (high press-
ure) opening plane
even with or above
nozzle entry plane
(Visually check for.
(both vertical and
(horizontal tip
(alignments;
(calibrated according
to Sec. 3.4.2
Repair or
return to
supplier
Differ-
ential *
pressure
gauge
Meets criteria
(Sec. 3.1.2);
agree, within 5%
of gauge-oil
manometer
(Check against a
(gauge-oil manometer
(at a minimum of 3
(points; 0.64
I(0.025); 12.7 (0.5);
(25.4 (1.0) ram (in)
JH20
I
Repair or
return to
supplier
Vacuum
gauge
0-760 mm (0-30
in.) Hg, ± 25 mm
((1 in.) at 380 mm
(in.) Hg
|Check against
(mercury U-tube
(manometer upon
|receipt
Adjust or
return to
supplier
(Continued)
-------
Section No. 3 .19.1
Date September 3. 1992 S~*\
Page 14 ( )
TABLE 1.1 (Continued)
Apparatus
Vacuum pump
Orifice
meter
Impingers
Filter
holder
(optional)
Filter
support
Filter
heating
system
Dry-gas
meter
|
1
(Acceptance limits
1
(Leak free; capable
|of maintaining a
| flow rate of
JO. 02 - 0.03 mVmin
j (0._66 to 1.1
(ftVmin) for pump
| inlet vacuum of
380 mm (15 in.) Hg
1
|AH@ of 46.74 ±
J6.35 mm (1.84 ±
J0.25 in.) H20 at
J68 °F (not
(mandatory)
Four Greenburg-
| Smith connected in
|a series, leak-
|free, noncontamin-
ating fittings
Leak-free;
jborosilicate glass
1
(Rigid stainless-
| steel wire screen
| Capable of
(maintaining filter
(holder at
temperature of
120 °C ± 14 °C
(248 °F ± 25°F)
Capable of
measuring volume
within 2% at a
flow rate of
0.02 mVmin
(0.75 ftVmin)
Frequency & method
of measurement
Check upon receipt
for leaks and
capacity
Upon receipt,
visually check for
damage and calibrate
against wet-test
meter .-,—
'. -
Visually check upon
receipt; check
pressure drop
(Subsec. 1.1.6)
Visually check
before use; conduct
leak check
Visually check upon
receipt, conduct
leak check
Visually check upon
receipt and run
heating system
checkout
Check for damage
upon receipt and
calibrate (Sec.
3.4.2) against
wet-test meter
Action if
requirements
are not met
Repair or
return to
supplier
Repair, if
possible,
otherwise
return to
supplier
Return to
supplier
As above
Repair or
return to
manufacturer
1
| Repair or
return to
| manufacturer
1
1
(Reject if
(damaged,
(behaves
erratically,
| or cannot be
(properly
adjusted t i
o
o
(Continued)
-------
TABLE 1.1 (Continued)
Section No. 3.19.1
Date September 3, 1992
Page 15
Apparatus
(Acceptance limits
Frequency & method
of measurement
(Action if
requirements
are not met
Acid Trap
(Mine Safety Appli-
|,ances air line
[filter acid ab-
|sorbing cartridge
Visually check upon
receipt
Return to
supplier
Thermo-
meters
|± 1 °C (2 °F) of
|true value in the
|range of 0 to
J25 °C (32 to
|77 °F) for impin-
|ger thermometer
land ± 3 °C
1(5.4 °F) of true
lvalue in the range
(of 0 to 90 °C
I (32 to 194 °F) for
|dry-gas meter
I thermometers
Check upon receipt
for dents or bent
stem, and calibrate
(Sec. 3.4.2) against
mercury-in-glass
thermometer
Reject if
unable to
calibrate
Barometer
(Capable of
(measuring
|atmospheric
(pressure within
(2.5 mm (0.1 in.)
JHg :
Check against a
mercury-in-glass
barometer or
equivalent ;••
calibrate
(Sec. 3.1.2)
Determine;
correction
factor, or
reject if
(difference
(more than
± 2.5 mm.
(0.1 in.) Hg
Gas density
determi-
nation
equipment
(Meet the
|requirements in
Sec. 3.2.1
Conduct checks shown
in Sec. 3.2.1,
upon receipt
Repair, ...-'"
replace, or
return to
supplier
Sample
Recovery
Glass
sample
bottles
(Leak-free, Tef-
|Ion lined caps,
1000 and 100 mL
Visually.check upon
receipt for cracks,
ensure that caps are
Teflon
Replace, or
return to. }
supplier
(Continued)
-------
Section No. 3 .19.1
Date September 3, 1992
Page 16
o
TABLE 1.1 (Continued)
Apparatus
Sample
Preparation
and
Analysis
Glassware
AA spec-
trometer
Recorder or
electronic
integrator
Optical
cell
Aeration
cell
Moisture
removal
system
Acceptance limits
. .' r
' ' •*
"• ,"'.
Class A
Suitable optical
resolution system
and detector
See owner's manual
."•'
See Figure 1.2
. •
See Figure 1.3
.
Heated cell or
moisture trap
to remove
condensation
'
Frequency & method
of measurement -
Visually check upon
receipt
Perform appropriate
calibrations
according to Sec. 5
'_
Upon receipt, check
"
Upon receipt, check
to ensure correct
dimensions, check
heating system
Visually check
Calibrate whenever
system is turned on
Action if
requirements
are not met
.
Replace or
return to
supplier
Return to
manufacturer
or repair and
re-check
Repair or
return to
manufacturer
Return, to
manufacturer ,
clean as
needed
1
Repair or
return to
manufacturer
[Calibrate
| heated cell
|or change
desiccant
Flowmeter
Capable of
measuring flow
of 1.5 L/min
(Calibrate with
|bubble meter or
|wet-test meter
|upon receipt
|Return to
| manufacturer
| or repair and.
I recalibrate
O
system
Regulator
to remove
condensation
from ,optical cell
Proper fittings
and pressure
control
|or change
| desiccant
1
Upon receipt, | Return to
attach to cylinder (manufacturer,
and check | repair,
j or replace
| fitting and
| re-check
!
O
(Continued)
-------
Section No. 3.19.1
Date September 3, 1992
Page 17
TABLE 1.1 (Continued)
Apparatus
Variable
transformer
Aeration
gas
cylinder
Tubing
Trip
balance
Analytical
balance
)
'Alternative
analytical
apparatus
Sampling
and
Sample
Recovery
Reagents
1
1
| Acceptance limits
1
(Capable of varying
(voltage from 0 to
(40 volts
1
(Nitrogen or dry,
|Hg-free air equip-
(ped with regulator
1
(See Sec. 1.3.13
(for specifications
(of tubing for the
| connections
1
(Capable of
(measuring within
|0.5 g
1
(Capable of weigh-
( ing to ± 0.5 mg
1
(Capable of gene-
| rating a linear
(calibration curve;
(two consecutive
(samples of equal
(size and concen-
tration agree ± 3%
(of average; and S
| 95% recovery of
| known concentra-
tion of spiked
| sample
I
1
1
| ACS reagent grade
| or Hg blank level
| specified
Frequency & method
of measurement
Visually check
upon receipt
Visually check
upon receipt
Visually check to
ensure proper type
tubing
Check with standard
weights upon receipt
and before each use
As above
See owner ' s manual
Visually check upon
receipt or conduct
Hg analysis
Action if
requirements
are not met
Return to
manufacturer
or repair
Return to
supplier
Replace
Replace or
return to
manufacturer
As above
Return to
supplier
Return to
supplier or
replace
(Continued)
-------
Section No. 3.19,
Date September 3,
Page 18
1952
TABLE 1.1 (Continued)
o
Apparatus
Water
Silica gel
Filter
(optional)
Analysis
Reagents
Filter
1
1
| Acceptance limits
1
(Deionized, dis-
| tilled meeting
IASTM 01193-77
I specifications
1
(indicating type,
| 6- to 16-mesh
1
(Glass fiber with-
|out organic bin-
(der; 99.95% col-
| lection efficiency
(for 0.3 urn dioctyl
Iphthalate smoke
| particles
1
(ACS reagent grade
(or equivalent;
(prepared as
(described in
(Sec. 1.5.3
1
1
1
(Whatman 40
| or equivalent
1
1
1
(Frequency & method
(of measurement
1
(check each lot or
(specify type when
(ordering
1
1
(Upon receipt, check
| label for grade
(or certification
1
| Manufacturer ' s gua-
(rantee that filters
(were tested accord-
| ing to -ASTM D 2986-
(71; observe under
(light for defects
1
1
1
(Upon receipt, check
| label for grade or
(certification; Check
(stability of
(prepared solution
(and prepare when
(necessary
1
(Upon receipt, check
| label for grade
1
Action if
requirements
are not met
Replace or
return to
supplier
Return to
supplier
Return to
supplier
Replace or
return to
supplier
Replace or
return to
supplier
O
o
-------
Section No. 3.19.2
Date September 3, 1992
Page 1
2.0
CALIBRATION OF APPARATUS
Calibrating the apparatus is one of the more important functions in
maintaining data quality. The detailed calibration procedures for the sampling
apparatus included in this section were designed for the sampling equipment specified
in Method 5 and described in the previous section. Calibrating the analytical
equipment is described in Section 3.19.5, which details the analytical procedures.
Table 2.1, at the end of this section, summarizes the quality assurance (QA) functions
for the calibrations. All calibrations, including those performed on the analytical
equipment} should 'be recorded on standardized forms and retained in a calibration log
book.
2.1
Metering System
The dry-gas meter (DGM) in the sampling system's meter console must be
calibrated against a primary standard meter (wet-test meter or spirometer). An
alternate procedure is to calibrate against a second reference meter (dry-gas meter or
critical orifice) that has been calibrated against a primary standard meter.
2.1.1 Wet-Test Meter—Wet-test meters are calibrated by the manufacturer to an
accuracy of ± 0.5%. The calibration must be checked initially upon receipt and yearly
thereafter. A wet-test meter with a capacity of 3.4vm3/h (120 ft3/h) or 30 L/revolution
(1 ftVrev) will be needed to calibrate the dry-gas meter. For large wet-test meters
(>30 L/rev), there is no convenient method for checking the calibration; consequently,
several methods are suggested, and other methods may be approved by the Administrator.
The initial calibration may be checked by any of the following methods:
1. Certification from the manufacturer that the wet-test meter is withir.
1% of true value at the wet-test meter discharge, so that only a leak
check of the system is then required.
2. Calibration by any primary-air or liquid-displacement method that
displaces at least one complete revolution of the wet-test meter.
3. Comparison against a smaller wet-test meter that has previously beer.
calibrated against a primary-air or liquid-displacement method, as
described in Section 3.5.2 of this Handbook.
4. Comparison against a dry-gas meter that has previously been calibrated
against a primary-air or liquid-displacement method.
The test-meter calibration should be checked annually. The calibration check
can be made by the same method as that of the original calibration; however, the
comparison method need not be recalibrated if the calibration check is within 1% of the
true value. When this agreement is not obtained, the comparison method or wet-test
meter must be recalibrated against a primary-air or liquid-displacement method.
2.1.2 Dry-Gas Meter as a Calibration Standard—A DGM may be used as a calibration
standard for volume measurements in place of the wet-test meter specified in Section
5.3 of Method 5, provided that it is calibrated initially and recalibrated periodically
as follows:
Standard Dry-Gas Meter Calibration—The DGM to be calibrated and used as a
secondary reference meter should be of high quality and should have appropriate
capacity (e.g., 3 L/rev [0.1 ftVrev] ) . A spirometer (400 L or more capacity), or
equivalent, may be used for this calibration, although a wet-test meter is usually more
practical. The wet-test meter should have a capacity of 30 L/rev (1 ftVrev) and
should be capable of measuring volume to within 1.0%. Wet-test meters should be
-------
Section No. 3.19.2
Date September 3, 19
Page 2
o
checked against a spirometer or a liquid displacement meter to ensure accuracy.
Spire-meters or wet-test meters of other sizes may be used, provided that the specified
accuracies of the procedure are maintained. The initial calibration may be checked by
any of the following methods:
1. Set up the components as shown in Figure 2.1. A spirometer, or
equivalent, may be used in place of the wet-test meter in the system.
2. Run the pump for at least 5 min at a flow rate of about 10 L/min (0.35
cfm) to condition the interior surface of the wet-test meter. The
pressure drop indicated by the manometer at the inlet side of the DGM
should be minimized (no greater than 100 mm H2O [4 in. H20] at a flow
rate of 30 L/min [1 cfm]). Using large diameter tubing connections and
straight pipe fittings will accomplish this minimization.
3. Collect the data as shown in the example data sheet (see Figure 2.2).
Make triplicate runs at each of the flow rates and at no less than five
different flow rates. The range of flow rates should be between 10 and
34 L/min (0.35 and 1.2 cfm) or over the expected operating range.
4. Calculate flow rate, Q, for each run using the wet-test meter volume
(Equation 2-1), Vu, and the run time, 9. Calculate the DGM coefficient
(Equation 2-2), Yd., for each run. These calculations are as follows:
~T—
P V
* bar w • -
Q = KI Equation 2-1
(ttt + t8td) 6
(tds + t6td) Pt
O
Equation 2-2
Vdfi (tw + tetd) (Pbar * Ap/13.6)
where :
Kj = 0.3858 for international system of units (SI); 17.64 for English
units.
Vw = Wet-test meter volume, liter (ft3).
Vds = Dry-gas meter volume, liter (ft3) .
tda = Average dry-gas meter temperature, °C (°F) .
tatd = 273 °C for SI units; 460 °F for English units.
tw = Average wet-test meter temperature, °C (°F) .
Pbir = Barometric pressure, mm Hg (in. Hg) .
Ap = Dry-gas meter inlet differential pressure, mm H2O (in. H2O) .
6 = Run time, min.
O
s* ">
-------
AIR IRIET
RSROnETER
\
V O CO
P P ra
(Q rt n
fl> (B ft
M-
W CO 0
(I) o
Figure 2.1. Sample meter system calibration setup.
vr>
»r>
M
-------
Date:
Dry-gas Meter Identification:
Barometric Pressure (P, ) :
_in. Hg
Approx-
imate
E low
rate
(0)
c£m
0.40
0.60
0.80
1.00
1.20
r— — i
Spiro-
meter
(wet
meter)
gas
volume
(V.)
ft'
r
Dry-
gas
meter
volume
(V^)
ftj
.
r
Spiro-
meter
(wet
meter)
(t.)
op
Pemperatur
I
Inlet
(t,)
oF
es
Dry-gas met
Outlet
(t,)
op
•
sr
Aver-
age
Aver-
age
meter
coef-
ficient
O rt
H*
^^§
^ Z
ID
O
o
Ii"u|iM'''> 2.2 Dry-c|iTn motor en 1 i brat i on rlntn form.
o
o
in
K)
-------
Section No. 3.19.2
Date September 3, 1952
Page 5
5. Compare the three Yd. values at each of the flow rates and determine the
maximum and minimum values. The difference between the maximum and
minimum values at each flow rate should be no greater than 0.030. Extra
sets of triplicate runs may be made to complete this requirement. Ir.
addition, the meter coefficients should be between 0.95 and 1.05. If
these specifications cannot be met in three sets of successive
triplicate runs, the meter is not suitable as a calibration standard and
should not be used as such. If these specifications are met, average
the three Yds values at each flow rate resulting in five average meter
coefficients, Yd..
6. Prepare a curve of meter coefficient, Yd>, versus flow rate, Q, for the
DGM. This curve shall be used as a reference when the meter is used to
calibrate other DGM's and to determine whether recalibration is
required.
Standard Dry-Gas Meter Recalibratioit-Recalibrate the standard DGM against a.
wet-test meter or spirometer annually or after every 200 hours of operation, whichever
comes first. This requirement is valid provided the standard DGM is kept in a
laboratory and, if transported, cared for as any other laboratory instrument. Abuse
to the standard meter may cause a change in the \calibration and will require more
frequent recalibrations.
As an alternative to full recalibration, a two-point calibration check may be
made. Follow the same procedure and equipment arrangement as for a full recalibration,
but run the meter at only two flow rates (suggested rates are 14 and 28 L/min [0.5 arid
.0 cfm]). Calculate the meter coefficients for these two points and compare the
•alues with the meter calibration curve. If the two coefficients are within 1.5% cf
the calibration curve values at the same flow rates, the meter need not be recalibrated
until the next date for a recalibration check.
2.1.3 Critical Orifices as Calibration Standards-Critical orifices may be used as
calibration standards in place of the wet-test meter specified in Section 5.3 of Methcd
5, provided that they are selected, calibrated, and used as follows:
Selection of Critical Orifices—The procedure that follows describes the use
of hypodermic needles or stainless-steel needle tubing that have been found suitable
for use as critical orifices. Other materials and critical orifice designs may be
used, provided the orifices act as true critical orifices (i.e., a critical vacuum car.
be obtained, as described in Section 7.2.2.2.3 of Method 5). Select five critical
orifices of appropriate size to cover the range of flow rates between 10 and 34 L/n^.n
or the expected operating range. Two of the critical orifices should bracket the
expected operating range.
A minimum of three critical orifices will be needed to calibrate a Method 5
DGM; the other two critical orifices can serve as spares, providing better selecticr.
for bracketing the range of operating flow
-------
Section No. 3.19.2
Date September 3, 1992
Page 6
rates. The needle sizes and tubing lengths shown below give the following approximate
flow rates:
o
Flow rate,
Gauge /cm
12/7.6
12/10.2
13/2.5
13/5.1
13/7.6
13/10.2
L/min
32.56
30.02
25.77
23.50
22.37
20.67
Flow rate,
Gauge /cm
14/2.5
14/5.1
14/7.6
15/3.2
15/7.6
15/10.2
L/min
19.54
17.27
16.14
14.16
11.61
10.48
These needles can be adapted to a Method 5-type sampling train as follows:
Insert a serum bottle stopper, 13- by 20-mm (0.5-in. by 75-in.) sleeve type, into a
13-mm (0.5-in.) Swagelok™ quick-connect fitting. Insert the needle into the stopper,
as shown in Figure 2.3.
Initial Critical Orifice Calibration—The procedure described in this section
uses the Method 5 meter box configuration with a DGM, as described in Section 2.1.8 of
Method 5, to calibrate the critical orifices. Other schemes may be used, subject to
the approval of the Administrator. The critical orifices must be calibrated in the
same configuration as they will be used (i.e., there should be no connections to the
inlet of the orifice) .
Prior to calibrating the critical orifices, the dry-gas meter in the meter box
must be calibrated. Before calibrating the meter box, leak check the system as
follows:
1. Fully open the coarse adjust valve and completely close the bypass
valve.
2. Plug the inlet.
3. Turn on the pump and determine whether there is any leakage. The
leakage rate must be zero (i.e., no detectable movement of the DGM dial
must be seen for 1 min).
4. Check also for leakages in the portion of the sampling train between the
pump and the orifice meter. See Section 5.6 for the procedure; make any
corrections, if necessary. If leakage is detected, check for cracked
gaskets, loose fittings, worn O-rings, etc., and make the necessary
repairs.
O
o
-------
f
i
Section No. 3.19.2
Date September 3, 19S2
Page 7
J
r_-7=^
JV5
1
-
-
.
k
m ii
1 II
H V
T
1
4
r
l
CRITICAL SERUM MICK
ORIFICE STOPPER CONNECT
Critical orifice adaptation to Method 5-type metering system.
o
METER BOX
DDDD
\
CRITICAL ORIFICE
Apparatus setup.
Figure 2.3 Critical orifice and apparatus setup.
-------
Section No. 3.19.2
Date September 3, 1992
Page 8
After determining that the meter box is leak-free, calibrate it according to
the procedure given in Section 5.3. Make sure that the wet-test meter meets the
requirements stated in Subsection 2.1.1. Check the water level in the wet-test meter.
Record the DGM calibration factor, Y. The critical orifice is then calibrated as
follows:
1. Set up the apparatus as shown in Figure 2.3.
2. Allow a warm-up time of 15 min. This step is important to equilibrate
the temperature conditions through the DGM.
3. Leak check the system as described above. The leakage rate must be
zero.
4. Before calibrating the critical orifice, determine its suitability anc
the appropriate operating vacuum as follows- Turn on the pump, fully
open the coarse adjust valve, and adjust the bypass valve to give e.
vacuum reading corresponding to about half an atmospheric pressure.
Observe the meter box orifice manometer reading, AH. Slowly increase
the vacuum reading until the meter box orifice manometer shows a stable
reading. Record the critical vacuum for each orifice. Orifices that
do not reach a critical value must not be used.
5. Obtain the barometric pressure using a barometer as described in Section
2.1.9 of Method 5. Record the barometric pressure, P^,., in mm Hg (in.
Hg) .
6. Conduct duplicate runs at a vacuum of 25 to 50 ram Hg (1 to 2 in. He)
above the critical vacuum. The runs must be at least S minutes each.
The DGM volume readings must be in increments of complete revolutions
of the DGM. As a guideline, the times should not differ by more thar.
3.0 s (this includes allowance for changes in the DGM temperatures) to
achieve ± 0.5% in K'. Record the information listed in Figure 2.4.
7. Calculate K' using Equation 2-3.
K, Vffl Y
AH/13.6)
1/2
o
o
K'
Equation 2-3
K' = Critical orifice coefficient, [ (m3) (°K)1/2] / [ (mm Hg)
, (min)H[(ftJ) (°R)1/2) ] / [ (in. Hg)(min)]}.
where:
Tamb = Absolute ambient temperature, °K (°R).
Average the K' values. The individual K' values should not differ by more than
from the average.
± 0.5%
o
-------
Section No. 3.19.2
Date September 3, 1992
Page 9
Date
Train ID
Critical orifice K' factor
Dry-Gas Meter
Final reading
Initial reading
Difference, Vm
Inlet /outlet temperatures
Initial
Final
Avg. temperature, tn
Time , 6
Orifice man. rdg., A H
Bar. pressure, Pbar
Ambient temperature, tmb
Pump vacuum
Critical orifice ID
m3 (ft3)
m3 (ft3)
m3 (ft3)
°C
°C
°C
min/s
min
mm (in.) H2O
mm (in.) Hg
°C (°F)
inm (in.) Hg
m3 (ft3)
Run number
1 2
DGM cal. factor, Y
Figure 2.4. Data sheet for determining DGM Y factor.
-------
Section No. 3.19.2
Date September 3. 1992
Page 10
Using the Critics! Orifices &s Calibration Standards—The dry-gas meter is
calibrated using the critical orifices as the secondary standard as follows:
1. Record the barometric pressure.
2. Calibrate the metering system according to the procedure outlined in
Sections 7.2.2.2.1 to 7.2.2.2.5. Record the information listed in
Figure 2.5.
3. Calculate the standard volumes of air passed through the DGM and the
critical orifices and calculate the DGM calibration factor, Y, using the
equations below:
Vm,.td) = K, V,, [PUr 4- (AH/13.6)]/Tm Equation 2-4
Vcrt.td. = K' (Pb.r e)/TMbl/2 Equation 2-5
Y = Vcrlltdl/Vm(.,dl Equation 2-6
where:
vcri«id! = Volume of gas sample passed through the critical orifice,
corrected to standard conditions, dscm (dscf).
-~r~
K' = 0.3858 °K/mm Hg for metric units
= 17.64 °R/in. Hg for English units.
4. Average the DGM calibration values for each of the flow rates. The
calibration factor, Y, at each of the flow rates should not differ by
more than ± 2% from the average.
Recalibration of critical orifices—To determine the need for recalibrating the
critical orifices, compare the DGM Y factors obtained from two adjacent orifices each
time a DGM is calibrated. For example, when checking orifice 13/2.5, use orifices
12/10.2 and 13/5.1. If any critical orifice yields a DGM Y factor differing by more
than 2% from the others, recalibrate the critical orifice according to the initial
calibration procedures above.
2.1.4 Sample Meter System-The sample meter system—consisting of the pump, vacuum
gauge, valves, orifice meter, and dry-gas meter—should be calibrated by stringent
laboratory methods before it is used in the field. The calibration should be
re-checked after each field test series. This re-check is designed to provide testers
with a method that can be used more often and with less effort, to ensure that the
calibration has not changed. When the quick check indicates that the calibration
factor has changed, testers must again use the complete laboratory procedure to obtain
the new calibration factor. After recalibration, the metered sample volume must be
multiplied by either the initial or the recalibrated calibration factor-that is, the
one that yields the lower gas volume for each test run.
o
o
o
-------
Date
Train ID
Critical orifice ID
Dry-Gas Meter ."
\
s~
Final reading
Initial reading
Difference, Vm
Inlet/outlet temperatures
Initial
Final
Avg. temperature, tm
Time, 8
Orifice man. rdg., A H
Bar. pressure, P^
Ambient temperature, t
Pump vacuum
K' factor
Average
Section No. 3 .19.2
Date September 3, 1992
Page 11
DGM cal. factor
m3 (ft3)
m3 (ft3)
m3 (ft3)
°C (°F)
• ••*""*
°C (°F)
°C (°F)
min/s
min
mm (in.) H20
mm (in.) Hg
°C (°F)
-"V
mm (in'.) Hg
Run number
1 2
Figure 2.5. Data sheet for determining K' factor,
-------
o
o
Section "No. 3'Yl9.2
Date September 3, 1992
Page 12
Before calibrating the metering system for the first time, conduct a leak
check. The meter system should be leak-free. Both positive (pressure) and negative
(vacuum) leak checks should be performed. The following pressure lea'k check procedure
will check the metering system from the quick-connect inlet to the orifice outlet and
will check the orifice-inclined manometer: '• •'
1. Disconnect the orifice meter line from the downstream orifice pressure
tap (the one closest to the exhaust of the orifice); plug this tap
(Figure 2.1).
2. Vent to the atmosphere the negative side of the inclined manometer. If
the inclined manometer is equipped with a three-way valve, this step can
be performed by turning the valve on the negative side of the ori-
fice-inclined manometer to the vent position.
3. Place a one-hole rubber stopper with a tube through its hole into the
exit of the orifice; connect a piece of rubber or plastic tubing, as
shown in Figure 2.1.
4. Open the positive side of the orifice-inclined manometer to the
"reading" position; if the inclined manometer is equipped with a
three-way valve, this will be the line position.
5. Plug the inlet to the vacuum pump. If a quick-connect with a leak-free
check valve is used on the control module, the inlet will not have, to
be plugged.
6. Open the main valve and the bypass valve.
7. Blow into the tubing connected to the end of the orifice until a
pressure of 127 to 178 mm (5 to 7 in.) H20 has built up in the system.
8. Plug or crimp the tubing to maintain this pressure.
9. Observe the pressure reading for a 1-min period. No noticeable movement
in the manometer fluid level should occur. If the meter box has a leak,
a bubbling-type leak check solution may aid in locating it.
After the metering system is determined to be leak-free by the positive leak
check procedure, the vacuum system to and including the pump should be checked by
plugging the air inlet to the meter box. If a quick-connect with a'leak-free stopper
system is presently on the meter box, the inlet will not have to be plugged. Turn the
pump on, pull a vacuum within 7.5 cm (3 in.) Hg of absolute zero, and observe the
dry-gas meter. If the leakage exceeds 0.00015 mVmin (0.005 f tVmin), the leak(s) must
be found and minimized until the above specifications are satisfied.
Checking the meter system for leaks before initial calibration is not
mandatory, but it is recommended.
Note: For metering systems with diaphragm pumps, the normal leak check
procedure described above will not detect leakages within the pump. For these cases,
the following leak check procedure is suggested: Make a 10-min calibration run at
0.00057 m3/min (0.02 ftVmin); at the end of the run, take the difference between the
measured wet-test meter and the dry-gas meter volumes; divide the difference by 10 to
get the leak rate. The leak rate should not exceed 0.00057 m'/rain (0.02 ftVmin) .
Initial calibration-The dry-gas meter and the orifice meter can be calibrate/*~N
simultaneously and should be calibrated when first purchased and any time the posttesl )
check yields a Y outside the range of the calibration factor Y +0.05 Y. A calibrated
wet-test meter (of proper size, with -t-1% accuracy) should be used to calibrate the
dry-gas meter and the orifice meter. The dry-gas meter and the orifice meter should
be calibrated in the following manner:
-------
Section No. 3.19.2
Date September 3, 1992
Page 13
1. Before its initial use in the field, leak check the metering system.
Leaks, if present, must be eliminated before proceeding.
2. Assemble the apparatus, as shown in Figure 2.6, with the wet-test meter
replacing the probe and impingers-that is, with the outlet of the
wet-test meter connected to a needle valve that is connected to the
inlet side of the meter box.
3. Run the pump for 15 min with the orifice meter differential (AH) set at
12.7 mm (0.5 in.) H20 to allow the pump to warm up and to permit the
interior surface of the wet-test meter to be wetted.
4. Adjust the needle valve so that the vacuum gauge on the meter box is
between 50 and 100 mm (2 to 4 in.) Hg during calibration.
5. Collect the information required on the forms provided (Figure 2.7).
Sample volumes, as shown, should be used.
6. Calculate Y4 for each of the six runs, using the equation in Figure 2.7
under the Y, column, and record the results on the form in the space
provided.
7. Calculate the average Y (calibration factor) for the six runs using the
following equation:
Yl + Y2
Y3
Y4 + Y5
Y6
Equation 2-7
-------
Section No. 3.19.2
Date September 3, 1S52
Page 14
o
THERMOMETERS
CONTROL
UTUBE
MANOMETER
PUMP
DRY GAS METER
WET TEST METER
Figure 2.6. Equipment arrangement for dry-gas meter calibration.
O
_J
o
-------
Section No. 3.19.2
Date September 3, 1392
Page 15
Date
Meter box number
1
Barometric pressure, Ph = in. Ha Calibrated by
Ori-
fice
mano-
meter
set-
ting
(AH),
in. H2O
0.5
Gas volume
Wet-
test
meter
(Vw),
ft3
5
Dry-
gas
meter
(Vd),
ft3
1.0 10
.
1.5 10
2.0 10
3.0 10
4.0
10
Temperatures
Wet-
test
meter
(tv),
oF
Drv-aas meter
Inlet
-------
Section No. 3.19 .2
Date September 3, 1992
Page 16
Nomenclature:
Vw = Gas volume passing through the wet-test meter, ft3.
Vd = Gas volume passing through the dry-gas meter, ft3.
tw = Temperature of the gas in the wet-test meter, °F.
tdl = Temperature of the inlet gas of the dry-gas meter, °F.
tdo = Temperature of the outlet gas of the dry-gas meter, °F.
td = Average temperature of the gas in the dry-gas meter, obtained by the
average tdl and tdo , °F.
AH = Pressure differential across orifice, in. H20.
YJ = Ratio of accuracy of wet-test meter to dry-gas meter for each run.
Tolerance Y, = Y ± 0.02 Y.
Y = Average ratio of accuracy of wet-test meter to dry-gas meter for all
six runs. Tolerance Y = Y ± O.Ol Y.
AH@, = Orifice pressure differential at each flow rate that gives 0.75
ftVmin of air at standard conditions for each calibration run, }
of H2O. Tolerance = AH@ ± 0.15 (recommended).
o
o
AH@ = Average orifice pressure differential that gives 0.75 ftVmin of air
at standard conditions for all six runs, in. H20. Tolerance =1.84
± 0.25 (recommended).
6 = Time for each calibration run, min.
Pt. = Barometric pressure, in. Hg.
Figure 2.7. Dry-gas meter calibration data (English units, backside).
-------
10.
Section No. 3.19.2
Date September 3, 1992
Page 17
Record the average on Figure 2.7 in the space provided.
Clean, adjust, and recalibrate, or reject the dry-gas meter if one or
more values of Y fall outside the interval Y ± 0.02 Y. Otherwise, the
average Y is acceptable and should be used for future checks and
subsequent test runs.
Calculate AH@, for each of the six runs using the equation in Figure
2.7A or 2.7B under the AH@j column, and record on the form in the space
provided.
Calculate'the average AH0 for the six runs using the following equation:
AH@1
AH03 + AH@4 + AHS5 -t- AH@6
Equation 2-8
Record the average on Figure 2.7 in the space provided.
11. Adjust the orifice meter or reject it i'f'^H@i varies by more than ± 3.9
mm (0.15 in.) H20 over the range of 10 to 100 mm (0.4 to 4.0 in.) K50.
Otherwise, the average AHS is acceptable-arid should be used for sub-
sequent test runs. „ .... <, ....,
Posttest calibration check—After each field itest series, conduct a
metering-system calibration check, as specified in Subsection 2.1.4, except for the
following variations:
1. Three calibration runs at a single intermediate orifice meter setting
may be used with the vacuum set at a maximum value reached during the
test series. The single intermediate orifice meter setting should be
based on the previous field test. A valve must be inserted between the
wet-test meter and the inlet of the metering system to adjust the
vacuum. " " '" • --v • '...., ..,..,,_ ,__i
2. If a temperature-compensating dry-gas meter was used, the calibration
temperature meter must be within ± 6 °C (10.8 °F) of the average merer
temperature during the test series.
3. Use Figure 2.8 to record the required information.
If the calibration factor Y deviates by <5% from the initial calibration
factor Y, then the dry-gas meter volumes obtained during the test series are
acceptable. If Y deviates by >5%, recalibrate the metering system and use whichever
meter coefficient (initial or recalibrated) yields the lowest gas volume for each test
run.
Alternate procedures (e.g., using the orifice meter coefficients or critical
orifices) may be used.
-------
Section No. 3 .19.2
Date September 3, 1992
Page 18
o
Date
Metering System ID No.
Barometric pressure, Pb =
Ori-
fice
mano-
meter
set-
ting
AH
in. Hg
'Spiro-
meter
(wet
.
Temperatures
| Spiro-
test) | Dry-gas
gas
meter
meter 1 Dry-qas meter
(wet |
volume (volume (meter)
-------
2.2
Section No. 3.19.2
Date September 3, 1992
Page 19
Temperature Gauges
2.2.1 Impinger Thermometer—The thermometer used to measure the temperature of the
gas leaving the impinger train should initially be compared with a mercury-in-glass
thermometer that meets ASTM E-l No. 63C or 63F specifications. This procedure is as
follows:
1. Place both the reference thermometer and the test thermometer in an ice
bath. Compare readings after they stabilize.
^- 2. Remove the thermometers from the bath and allow both to come to room
temperature. Again, . compare readings after they stabilize.
3. Accept the test thermometer if its reading agrees to within 1 °C (2 °F)
of the reference thermometer reading at both temperatures. If the
difference is greater than 1 °C (2 °F), the thermometer should be
adjusted and recalibrated until the criteria are met, or it should be
rejected. Record the results on Figure 3.1 of Section 3.19.3.
2.2.2 Dry-gas Thermometers—The thermometers used to measure the metered gas sample
temperature should be compared initially with a mercury-inglass thermometer as above,
using a similar procedure.
1. Place the dial type (or equivalent1}" thermometer and the mercury-in-glass
thermometer in a hot water bath, 40 to 50 °C (104 to 122 °F). Compare
the readings after they stabilize.
2. Allow both thermometers to come to room temperature. Compare readings
after thermometers stabilize.
3. Users should accept the dial type (or equivalent) thermometer under the
following conditions: The values must agree to within 3 °C (5.4 °F) at
both points; the temperature differentials at both points are within 3
°C (5.4 °F), and the temperature differential is taped to the thermome-
ter and recorded on the pretest sampling check form (Figure 3.1).
4. Prior to each field trip, compare the temperature reading of the
mercury-in-glass thermometer at room temperature with that of the meter
system thermometer. The values or corrected values should be within 6
°C (10.8 °F) of one another, or the meter thermometer should be replaced
or recalibrated. Record any temperature correction factors on Figure
3.1 of Section 3.19.3 or on a similar form.
2.2.3 Stack Temperature Sensor—The stack temperature sensor should be calibrated
upon receipt or checked before field use. Each sensor should be uniquely marked for
identification. The calibration should be performed at three points and then extra-
polated over the range of temperatures anticipated during actual sampling. For the
three-point calibration, a reference ASTM mercury-in-glass thermometer should be used.
The following procedure is recommended for calibrating stack temperature
sensors (thermocouples and thermometers) for field use.
I. For the ice-point calibration, form a slush from crushed ice and liquid
water (preferably deionized, distilled) in an insulated vessel such as
a Dewar flask. Taking care that they do not touch the sides of the
flask, insert the stack temperature sensors into the slush to a depth
of at least 2 in. Wait 1 min to achieve thermal equilibrium and record
the readout on the potentiometer. Obtain three readings taken at i-min
intervals.
Note: Longer times may be required to attain thermal equilibrium
with thick-sheathed thermocouples.
-------
Section No. 3 .19 .2
Date September 3,
Page 20
^™^
V J
Fill a large Pyrex beaker with water to a depth >4 in. Place several
boiling chips in the water and bring the water to a full boil using a
hot plate as the heat source.
Insert the stack temperature sensor (s) in the boiling water to a
depth of at least 2 in., taking care not to touch the sides or bottom
of the beaker.
Place an ASTM reference thermometer alongside the sensor (s) . If
the entire length of the mercury shaft in the thermometer cannot be
immersed, a temperature correction './ill be required to give the correct
reference temperature.
After 3 min, both instruments will attain thermal equilibrium.
Simultaneously record temperatures from the ASTM reference thermometer
and the stack temperature sensor three times at 1-min intervals.
For thermocouple, repeat Step 2 with a liquid (such as cooking oil) that
has a boiling point in the 150 to 250 °C (300 to 500 °F) range. Record
all data on Figure 2.9. For thermometers other than thermocouples,
repeat Step 2 with a liquid that boils at the maximum temperature at
which the thermometer is to be used, or place the stack thermometer and
reference thermometer in a furnace or other device to reach the required
temperature . -«~
Note: If the thermometer is to be used at temperatures higher than the
reference thermometers can record, the stack thermometer may be
calibrated with a thermocouple previously calibrated with the above^—. .
procedure . { )
If the absolute values of the reference thermometer and thermocouple (sV_X
agree to within 1.5% at each of the three calibration points, plot the
data on linear graph paper and draw the best-fit line to the three
points or calculate the constants of the linear equation using the
least -square method. The data may be extrapolated above and below the
calibration points to cover the entire manufacturer's suggested range
for the thermocouple. For the portion of the plot or equation th&-
agrees within 1.5% of the absolute reference temperature, no correcticr.
need be made. For all portions that do not agree within 1.5%, use the
plot or equation to correct the data.
If the absolute values of the reference thermometer and stack
temperature sensor (other than the thermocouple) agree to within 1.5%
at each of the three points, the thermometer may be used over the ranee
of calibration points for testing without applying any correcticr.
factor. The data cannot be extrapolated outside the calibration points.
-------
Date
Ambient temperature
Calibration person
Thermocouple No.
°F Barometric pressure
Reference: mercury-in-glass
other
Section No. 3.19.2
Date September 3, 1992
Page 21
in. Hg
op
oF
Reference
point
number
|Source"
j(specify)
I
(Reference
I thermometer
|temperature,
OF
I
I Thermocouple
|potentiometer
I temperature,
I Temperature*
I difference,
1%
Type of calibration system used.
(ref temp, °F + 460) - (test thermom temp, °F + 460) x 10Q _ (
-------
Section No. 3.19 .2
Date September 3, 19S2
Page 22
o
2.3 Probe Heater
The probe heating system should be calibrated prior to field use according to
the procedure outlined in APTD-0576. Probes constructed according to APTD-0581 need
not be calibrated if the curves of APTD-0576 are used.
2.4 Barometer ..>•„, -
The field barometer should be adjusted initially' and before each test series:
to agree to within 2.5 mm (0.1 in.) Hg of the mercury-inglass barometer or with the
station pressure value reported by a nearby National Weather Service station, corrected
for elevation. The correction for elevation difference between the station and the
sampling point should be applied at a rate of -2.4 mm Hg/30 m (-0.1 in. Hg/100 ft).
Record the results on the pretest sampling check form (Figure 3.1 of Section 3.19.3).
2 .5 Probe Nozzle
Probe nozzles should be calibrated before initial use in the field. Using a
micrometer, measure the ID of the nozzle to the nearest 0.025 mm (0.001 in.). Make
three measurements using different diameters each time, and obtain the average. Tne
difference between the high and the low numbers should not exceed 0.1 mm (0.004 in.
When nozzles become nicked, dented, or corroded, they should be reshaped, sharpened,^
and recalibrated before use. Each nozzle should be permanently and uniquely
identified. Figure 2.10 is an example of a nozzle calibration data form.
2.6 Pitot Tube
The Type S pitot tube assembly should be calibrated using the procedure
outlined in Section 3.1.2 of this Handbook for Method 2.
2 .7 Trip Balance
The trip balance should be calibrated initially by using Class £ standard
weights and should be within 0.5 g of the standard weight. Adjust or return the
balance to the manufacturer if limits are not met.
O
-------
Section No. 3.19.2
Date September 3, 1992
Page 23
I/Date
Calibrated by
Nozzle
ID No.
Nozzle Diameter*
mm (in.)
1
mm (in.)
im (in.)
'
~f~
\
ADb
mm ( in . )
-
1
mm ( in . )
where :
8 EI.;, 3 = Three different nozzle diameters, mm (in. ; each diameter must be measured
AD
within (0.025 mm) 0.001 in.
Maximum difference between any two diameters, mm (in.), AD <(0.10 mm) 0,004
in.
Average of D,, D;, and D3.
Figure 2.10. Nozzle calibration data form.
-------
Section No. 3.19.2
Date September 3, 1992
Page 24
TABLE 2.1. ACTIVITY MATRIX FOR EQUIPMENT CALIBRATION
o
Apparatus
Wet-test
meter
Dry-gas
meter
Critical
Thermo-
meter
Probe
heating
system
Barometer
Acceptance limits
Capacity S3. 4
mVh (120 ftVh);
accuracy within
± 1.0%
Yj = Y ± 0.02 Y
K' = K i 0.03 K'
Impinger thermo-
meter ± 1 °C (2
°F) ; dry-gas meter
thermometer ± 3 °C
(5.4 °F) over
range ; stack
temperature sensor
± 1.5% of absolute
temperature
Capable of
maintaining 120 °C
± 14 °C (248° ± 25
°F) at a flow rate
of 20 L/min
(0.71 ft'/min
± 2.5 mm (0.1 in.)
Hg of mercury-in-
glass barometer
Frequency & method
of measurement
Calibrate initially,
and then yearly by
liquid displacement
Calibrate vs.
wet-test meter
initially, and when
posttest check
exceeds Y ± 0.05 Y
Calibrate vs. wet,
dry, or~ bubble meter
upon receipt and
after each test
Calibrate each ini-
tially as a separate
component against a
mercury-in-glass
thermometer; then
before each trip
compare each as part
of the train with
the mercury-in-glass
thermometer
Calibrate component
initially by APTD-
0576; if constructed
by APTD-0581, or use
published calibra-
tion curves
Calibrate initially
vs. mercury-in-glass
barometer; check
before and after
each field test
(Action if
| requirements
| are not met
1
(Adjust until
| specifications
(are met, or
jturn to
(manufacturer
1
(Repair, or
| replace and then
(recalibrate
1
1
(Repair and then
| recalibrate,
(or replace
1
(Adjust;
(determine
|a constant
(correction
| factor;
(or reject
1
1
1
(Repair, or
| replace and
(then reverify
(the calibration
1
1
(Adjust to agree
(with a certified
(barometer
1
1
O
(Continued)
o
-------
TABLE 2.1. (Continued)
Section No. 3.19.2
Date September 3, 1992
Page 25
Apparatus
I
(Acceptance limits
|Frequency & method
I of measurement
I Action if
|requirements
I are not met
Probe
nozzle
(Average of three
|ID measurements of
(nozzle; difference
(between high and
| low SO . 1 mm
I(0.004 in.)
|Use a micrometer to
(measure to nearest
(0.025 mm (0.001 in.)
I
I
(Recalibrate,
I reshape, and
I sharpen when
(nozzle becomes
(nicked, dented
|corroded
Trip
balance
|500-g capacity;
|capable of measur-
|ing within ± 0.5 g
(check with standard
(class S weights upon
(receipt
(Adjust, replace
(or return to
I manufacturer
-------
(f
Section No. 3.19.3
Date September 3, 1992
Page 3
3.1.7 Bare/neter-The field barometer should be compared with the mercury-in-glass
barometer or the weather station reading, after making an elevation correction, prior
to each field trip.
3 .2 _Sample Recovery Equipment and Reagents
Clean all sample exposed-glassware using the following procedures:
1. Soak glassware in 50% HN03 for a minimum of 1 h.
2. Rinse with tap water.
3. Rinse with 8 N HCl.
4. Rinse with tap water.
5. Rinse with DI water.
3.2.1 Glass Sample Bottles—The sample bottles must be leak-free, must gave
Teflon-lined caps, and must be 1000 and 100 mL in size.
3.2.2 Graduated Cylinder—A 250-mL graduated cylinder is required.
3.2.3 Funnel an'd Rubber Policeman—These items aid in transferring the silica gel to
the container; they are not necessary if the silica gel is weighed in the field.
I"
3.2.4 Funnel—A glass funnel is required to aid in sample recovery.
3.3 Equipment Packing ,
The accessibility, condition, and functioning of measurement devices in the,
field depend on packing them carefully and on moving them carefully at the site.
Equipment should be packed to withstand severe treatment during shipping and field
operations. The material used to construct shipping cases is therefore important. The
following containers are suggested, but they are not mandatory.
3.3.1 Proie—Seal the inlet and outlet of the probe to protect it from breakage and
pack it in the container. An ideal container is a wooden case (or the equivalent)
lined with foam material and having separate compartments to hold the individual
probes. The case should have handles or eye-hooks that can withstand hoisting and that
are rigid enough to prevent bending or twisting during shipping and handling.
3.3.2 Impingers, Connectors, and Assorted Glassware-All impingers and glassware
should be packed in rigid containers and protected by polyethylene or other suitable
material. Individual compartments for glassware will help to organize and protect each
piece.
3.3.3 Volumetric Glassware-A sturdy case lined with foam material can contain dryir.g
tubes and assorted volumetric glassware.
3.3.4 Meter Box—The meter box, which contains the manometers, orifice meter, vacuum
gauge, pump, dry-gas meter, and thermometers, should be packed in a shipping container
unless its housing is sufficient to protect components during travel. Additional puz.p
oil should be packed if oil is required. A spare meter box should be included in case
of failure.
3.3.5 Wash Bottles and Storage Containers-Storage containers and miscellaneous
glassware should be packed in -a rigid, foam-lined container.
-------
Section No. 3.19.3
Date September 3, 1992
Page 4
3.3.6 Chemicals-Chemicals should be packed in a rigid, foam-lined container.
As mentioned in Subsection 1.5.1.6 (Absorbing Solution, 4% KMn04) , caution
must be exercised for the storage and transport of KMnO<. To prevent autocatalytic
decomposition of the permanganate solution, filter it through Whatman 541 filter paper.
The reaction of the KMnO4 with the acid may cause pressure buildup in the sample
storage bottle. These bottles should not be filled to capacity and should be vented
to relieve excess pressure and to prevent explosion of the sample. A No. 70-72 hole
drilled in the container cap and Teflon liner is recommended.
Also, caution should be exercised with the HC1 reagent because it is highly
corrosive. . • '' ' ' .
' • ' • t. ••!":'':. j
j "
3.3.7 Safety Equipment for Sampling Train Preparation and Sample Recovery-Safety
glasses and protective laboratory gloves should be packed for the personnel assigned
to prepare the sampling train and recover the sample. Serious injury can result from
contact with HC1 and KMn04.
o
o
X
X
-------
Section No. 3.19.3
Date September 3, 1952
Page 5
TABLE 3.1 ACTIVITY MATRIX FOR PRESAMPLING OPERATIONS
Apparatus
Apparatus
Check and
Calibration
Nozzles
and
probe
liners
1
(Acceptance limits
1
1
1
1
1
|l. Probe heating
| system capable of
| heating to 120 °C
|± 14 °C at a flow
jrate of 20 L/min
1
| 2 . Probe leak free
jat 380-mm (15 in.)
(Hg
1
|3. Nozzles
(calibrated
| (Sec. 3.19.2
I Subsec . 2.4)
1
1
(Frequency & method
| of measurement
!
1
1
1
1
|l. Check heating
(system initially and
(when moisture cannot
|be prevented during
(testing
1
| 2 . Visually check
(before test
1
1
|3. Before -test to
(nearest O.D25-irun
(with micrometer
1
1
Action if
requirements
are not met
1 . Repair or
replace
2 . Replace
3. Recalibrate,
reshape, or
replace
(4. Probe and
(nozzle free of
|contaminants
j(Sec. 3.2)
I
14. Clean internally
(by brushing with tap
(water, deionized
(distilled water, and
(acetone; air dry
(before test
|4. Repeat
|cleaning and
|assembly
|procedures
I
Impingers,
filter
holders,
and other
glassware
(Meets specifica-
|tions in Subsec. 1
(of Sec. 3.19.1;
|cleaned according
|to Sec. 3.19.3
|Subsec. 3.1.2; and
(sealed with
(Teflon or glass
|stoppers
Before each test
(Repair or
(discard and
replace
Dry-gas
meter
(clean and readings
(within 2% of
|average
(calibration factor
(Calibrate according
(to Sec. 3.19.2
I
Repair or
replace and ther
recalibrate
Filters
|Free of
|irregularities
(visually check prior
(to testing
(Replace
(Continued)
-------
TABLE 3.1 (Continued)
Section No. 3.19.3
Date September 3, 1992
Page 6
o
Apparatus
I
(Acceptance limits
(Frequency & method
I of measurement
(Action if
[requirements
are not met
Silica gel
(indicating, 6-16
|mesh, use fresh-
|or dry-used silica
(gel at 175 °C
(350 °F)
I If moisture content
|is to be determined,
(weigh several 200-
jto 300-g portions of
(silica gel
11 (± 0.5 g); use
(airtight containers;
|record weight of
(container plus
(silica gel
(calibrated, within (Calibrate against
|mercury-in-glass •
|thermometer
I(Sec. 3.4.2) before
| each
Replace or
reweigh
Thermo-
meters
°C (2 °F) for
|impinger thermo-
lmeter, ± 3 °C
I (5.4 °F) for
(dry-gas meter
|thermometer
Replace
O
Barometer
(Calibrated, within
J2.5-mm (0.1 in.)
(Hg
(calibrate against
| mercury-in-glass
(barometer (Sec.
j 3 .7 .2) before each
test
Replace
Sample
Recovery
Equipment
and
Reagents
Glass
sample
bottles
Graduated
cylinder
(Clean, leakless,
(Teflon-lined caps
(clean, glass and
(class A; 250 mL
(with <2 mL
I subdivisions
(Before each field (Replace
(test I
I I
I I
(Before each field (Replace
|trip check for |
(cracks, breaks, and |
(manufacturer flaws |
Funnel
(Clean, glass,
(Class A
Same as above
Same as above
O
(Continued)
-------
Section No. 3.19.3
Date September 3, 1992
Page 7
TABLE 3.1 (Continued)
Apparatus
Equipment
packing
Probe
Impingers ,
connectors,
and
assorted
glassware
Volumetric
glassware
Meter box
Wash
bottles
and
storage
containers
Chemicals
Acceptance limits
Rigid container
protected by
polyethylene foam
Rigid container
protected by
polyethylene foam
Packed in original
containers, if
available, or a
rigid container
lined with foam
and marked
•Fragile"
_|
Meter box case
and/or additional
material to
protect train
components; pack
spare -meter box
Rigid foam-lined
container
Rigid foam-lined
container
I Frequency & method
of measurement
Prior to each
shipment
Prior to each
shipment
Prior -ifb each
shipment \
Prior to each
shipment
Prior to each
shipment
Prior to each
shipment
[Action if
requirements
are not met
Repack
Repack
Repack
Repack
Repack
Repack
Y
-------
o
o
o
-------
Section No. 3.19.4
Date September 3, 1992
Page 1
4.0 ON-SITE MEASUREMENTS
On-site activities include transporting the equipment to the test site, unpacking
and assembling the equipment, sampling for particulate and gaseous mercury, and
recording the data. The associated QA activities are summarized in Table 4.1 at the
end of this section.
4.1 Transport of Equipment to the Sampling Site
The most efficient means of transporting the equipment from ground level to the
sampling site (often above ground level) should be decided during the preliminary site
visit or by prior correspondence. Care should be taken to prevent damage to:the
equipment or injury to .test personnel during the moving. A clean "laboratory type"
area free of excessive dust and mercury' compounds should be located and designated for
preparing the nozzle, probe, filter holder, and impingers and for sample recovery.
4.2 Preliminary Measurements and Setup
A preliminary survey should be conducted prior to sampling and analysis, unless
adequate prior knowledge of the source is available. Testing must be conducted at the
proper sampling locations and during the proper process and control equipment operating
cycles or periods. Testers should refer to Subsection 3.19.3.1 for information
ypically needed to establish the proper sampling and analysis protocol.
Testers should have calculated the minimum sampling run time required, unless it
is known that the minimum time stated by the applicable regulations will be sufficient
to provide proof of compliance.
In this method, highly oxidizable matter may make it impossible to sample for the
desired minimum time. This problem is indicated by the complete bleaching of the
purple color of the KMnO4 solution. In these cases, testers may divide the sample run
into two or more subruns to ensure that the absorbing solution will not be depleted.
In cases where excess water condensation is encountered, collect two runs to make one
sample. •••.•-.,
4.2.1 Preliminary Measurements and Setup-The sampling site should be selected in
accordance with Method 1. If the duct configuration or some other factor makes this
impossible, the site should be approved by the Administrator, prior',to conducting the
test. A 115-V, 30-A electrical supply is necessary to operate the standard sampling
train. Either measure the stack and determine the minimum number of.;traverse points
by Method 1, or check the traverse points determined during the preliminary site visit
(Section 3.0) . Record all data on the traverse point location,form shown in Method 1.
These measurements will be used to locate the pitot tube and the sampling probe during
preliminary measurements and actual sampling.
-------
Section No. 3.19,
Date September 3,
Page 2
1992
o
4.3 Preparations for Sampling
The most common situations and problems are addressed in this section. Both
required and recommended QA/control checks and procedures are provided to assist in
collecting data of acceptable quality and to assess the accuracy of the sampling and
analysis.
On-site sampling includes the following steps:
1. Conducting preliminary measurements and setting up the recovery area.
2. Preparing and setting up the sampling system for leaks.
3. Connecting electrical service and checking the sampling system for leaks.
4. Heating the probe and filter to the proper temperature.
5. Inserting the probe into the duct and sealing the duct.
6. Isokinetic sampling.
7. Recording data.
8. Posttest leak check of the sampling system.
9. Recovering the sample and transporting it to the laboratory.
4.3.1 Stack Parameters-Check the sampling site for cyclonic or nonparallel flow as
described in Method 1 (Section 3.0). The sampling site must be acceptable before a
valid sample can be taken. Determine the stack~rpressure, temperature, and the range
of velocity heads encountered (Method 2) . Determine the moisture content using the
approximation Method 4, or its alternatives, for the purpose of setting the isokinetic
sampling rate. If the identical source has been tested before or if a good estimateX^X
of the moisture content is available, this should be sufficient. The reference methodt )
(Section 3.4.10) uses the condensate collected during sampling to determine the
moisture content used in final calculations. If the stack is saturated with moisture
or has water droplets, the moisture content must also be determined by partial pressure
with the use of a more accurate stack gas temperature sensor (Method 4).
Determine the dry molecular weight of the stack gas, as required in Method 2.
If an integrated gas sample is required, follow Method 3 procedures and take the gas
sample simultaneously with, and for the same total length of time, as the particulate
run. The sampling and the analytical data forms for molecular weight determinations
are in Method 3.
Using the stack parameters obtained by these preliminary measurements, the tester
can set up the nomograph as outlined in APTD-0576 or use a calculator. An example of
a nomograph data form is.shown in Figure 4.1 of the Method 5, Section 3.4.4.
Select a nozzle size based on the range of velocity heads, so that it is not
necessary to change the size to maintain isokinetic sampling rates during thev>run.
Install the selected nozzle using a Viton A O-ring when glass liners are used. Other
connecting systems such as Teflon ferrules may be used. Mark the probe with heat
resistant tape or by some other acceptable method to denote the proper distance into
the stack or duct for each sampling point. Select a total sampling time greater than
or equal to the minimum total sampling time specified in the test procedures for the
specific industry so that:
1. The sampling time per point is >2 min (a greater time interval may be
specified by the Administrator).
2. The sample volume corrected to standard conditions exceeds the required
minimum total gas sample volume.
The latter can be based on an approximate average sampling rate. Its-
recommended that the number of minutes sampled at each point be either an integer o"
an integer plus one-half minute to avoid timekeeping errors. In some circumstances
(e.g., batch cycles), it may be necessary to sample for shorter times- at the traverse
D
-------
Section No. 3.19 .4
Date September 3, 1992
Page 3
points and to obtain smaller gas sample volumes. In these cases, the Administrator's
approval must be obtained first.
4.3.2 Sampling Train Preparation—During preparation of the sampling train, keep all
openings where contamination can occur covered until just prior to assembly or until
sampling commences. The glassware should have been cleaned as described in Section
3.19.3 by soaking in 50% HNO, and then rinsing with tap water, 8 N HC1, tap water, and
finally deionized distilled water. Prepare the individual sampling train components
as follows:
Impinqers
1. Place 50 ml of fresh 4% KMnO4 in the first cleaned impinger using a graduated
cylinder that has been properly cleaned,
2. Place 100 ml of fresh 4% KMn04 in the second and third impingers using a
graduated cylinder, and
3. Place 200 to 300 g of preweighed silica gel in the fourth impinger.
Precaution: It is extremely important that all.sample recovery personnel wear
safety glasses and gloves due to the dangers associated with impinger solutions and
recovery solutions.
Record the weight of the silica gel and the container on the sample recovery data
form, Figure 4.1, or other similar data form. -'"Place the empty container in a safe
place for use later in the sample recovery. If moisture content is to be determined
by impinger analysis, weigh each of the first three impingers to the nearest 0.5 g, and
record these weights. Place the silica gel container in a clean place for later use
in the sample recovery. Alternatively, the weight of the silica gel plus impinger may
be determined to the nearest 0.5 g and recorded.
-------
Section No. 3.19 .4
Date September 3, 1992
Page 4
o
Plant Sample Data
Sample Location Run No.
Sample Recovery Person Recovery Date
Filter(s) No.
MOISTURE
Impinqers
Final volume (wt) ml (g) Final wt g g
Initial volume (wt) ml (g) Initial wt g g
Net volume (wt) ml (g) Net wt g g
Total moisture g
RECOVERED SAMPLE BLANK
Blank filter Container No. KMnO4 added, sealed and level marked?
Blank KMnO4 solution (650 mL) Container No. Sealed and level marked?
Blank HC1 solution (25 mL added to 200 mL H2O) Container No.
Sealed and level marked?
RECOVERED SAMPLE
KMn04 impinger contents and rinse (400 mL) Container No.
Sealed and level marked?
Filter Container No. KMnO4 added, sealed and level marked?
HCl solution (25 mL added to 200 mL H,O) Container No.
Sealed and level marked?
Samples stored and locked?
Remarks:
O
Date of laboratory custody
Laboratory person taking custody
Remarks:
Figure 4.1. Sample recovery and integrity data form.
-------
Section No. 3.19.4
Date September 3, 1992
Page 5
The use of a filter is optional in Method 101A. However, because of the
digestion techniques used for sample preparation, it is highly recommended that a
falter be used. Assemble the filter holder as follows:
Filter (optional)
1. Using a tweezer or clean disposable surgical gloves, place a filter in the
filter holder. Be sure that the filter is properly centered and that the
gasket is properly placed to prevent the sample gas stream from circumventing
the filter.
2. Visually check the filter for damage after the assembly is completed.
3. The filter or filter sample container should be marked.
Record the filter number on the sample recovery data form and then place the
filter sample container in a clean place for later use in the sample recovery.
Assemble the probe and nozzle as follows:
Probe/nozzle assembly
1. The probe liner should be glass and cleaned using the procedures described
above.
2. Place the properly sized, calibrated, and cleaned nozzle on the inlet of the
probe using a Teflon ferrule or Viton O-ring connection.
The nozzle should be uniquely identified. Record the nozzle number and diameter
on the sampling data form. -*~
\
4.3.3 Sampling Train Assembly—Assemble the train'as shown in Figure 1.1, using (if
necessary) a very light coat of silicone grease only on the outside of all ground-glass
joints to avoid contamination. The tester may find that it is beneficial to conduct
a leak check of the. sampling train in the assembly area prior to taking the system to
the;stack.
The sampling train is then transported to the stack. At the stack, place crushed
ice and water around the impingers. If not already an integral part of the probe
assembly, a temperature sensor should be attached to the metal sheath of the sampling
probe so that the sensor extends beyond the probe tip and does not touch any metal.
The sensor's position should be about 1.9 to 2.54 cm (0.75 to 1 in) from the pitot tube
and the nozzle to avoid interference with the gas flow. Alternative arrangements are
shown in Method 2.
4.3.4 Sampling Train Leak Checks—Leak checks are necessary to assure that the sample
has not been biased low by dilution air. The reference method (Section 3.19.10)
specifies that leak checks be performed at certain times as discussed below.
Pretest—A pretest leak check is recommended, but not required. If the tester
opts to conduct the pretest leak check, follow the procedure described below:
After the sampling train has been assembled, set the filter heating system at the
desired operating temperature. Allow time for the temperature to stabilize. If a
Viton A O-ring or other leak free gasket is used in connecting the probe nozzle to the
probe liner, leak check the train at the sampling site by plugging the nozzle and
pulling a 380- mm (15 in) Hg vacuum. Note: A lower vacuum may be used if it is not
exceeded during the test.
If an asbestos string is used for the probe gasket, do not connect the probe to
the train during the leak check. Instead, leak check the train by first plugging the
inlet to the filter holder and pulling a 380-mm (15 in) Hg vacuum (see note in the
previous paragraph) . Then connect the probe to the train and leak check at about 25-iran
(1 in.) Hg vacuum; alternatively, the probe may be leak checked with the rest of the
sampling train in one step at a 380-mm (15 in.) Hg vacuum. Leakage rates >4% of the
average sampling rate or 0.00057 mVmin (0.02 ftVmin) , whichever is less, are
unacceptable.
-------
Section No. 3.19.4
Date September 3, 1992
Page 6
The following leak check instructions for the sampling train are taken from APTD- \)
05813 and APTD-0576. Start the pump with the bypass valve fully open and the coarse
adjust valve closed. Open the coarse adjust valve and then slowly close the bypass
valve until the desired vacuum is reached. Do not reverse the direction of the bypass
valve; this will cause KMnO4 solution to back up from the impingers into the filter
holder. If the desired vacuum is exceeded, either leak check at this higher vacuum or
end the leak check as described below and start over.
When the leak check is complete, first slowly remove the plug from the inlet to
the probe or the filter holder and then close the coarse adjust valve and immediately
turn off the vacuum pump. (This prevents the KMnO4 in the impingers from being forced
back into the filter holder and prevents the silica gel from being forced back into the
third impinger.) Visually check to be sure KMn04 did not contact the filter and that
the filter has no tears before beginning the sampling.
During the Sampling—If. a component (e.g., filter assembly or impinger) change is
necessary during the sampling run, a leak-check should be conducted before the change.
The leak-check should be done according to the procedure outlined above, except that
it should be at a vacuum equal to or greater than the maximum value recorded up to that
point in the test. If the leakage rate is <0.00057 mVmin (0.02 ftVmin) or 4% of the
average sampling rate (whichever is less), the results are acceptable, and no
correction need be applied to the total volume>~of dry gas metered. If, however, a
higher leakage rate is obtained, the tester either>should record the leakage rate and
plan to correct the sample volume as shown in Section 6.3(b) of the Reference Method
(Section 3.19.10), or should void the sampling run. Note: Be sure to record the dry
gas meter reading before and after each leak-check performed during and after each test
run so that the sample volume can be corrected.
Posttest—A leak-check is mandatory at the conclusion of each sampling run. The
leak-check should be in accordance with the procedures in this section and at a vacuum
equal to or greater than the maximum value reached during the sampling run. If the
leakage rate is <0.00057 mVmin (0.02 ftVmin) or 4% of the average sampling rate
(whichever is less), the results are acceptable, and no correction need be applied to
the total volume of dry gas metered. If, however, a higher leakage rate is obtained,
the tester either should record the leakage rate and correct the sample volume as shown
in Section 6.3(a) or 6.3(b) of the Reference Method (Section 3.19.10), or should void
the sample run. Note: Be sure to record the dry gas meter reading before and after
performing the leak check so that the sample volume can be corrected.
4.3.5 Sampling Train Operation—Just prior to sampling, clean the portholes to minimize
the chance of sampling deposited material. Verify that the probe and the filter
heating systems are up to the desired temperatures and that the pitot tube and the
nozzle are located properly. Follow the procedures below for sampling.
1. Record the initial dry gas meter readings, barometric pressure, and other
data as indicated in Figure 4.2.
2. Position the tip of the probe at the first sampling point with the nozzle tip
pointing directly into the gas stream. When in position, block off the open
area around the probe and the porthole to prevent flow disturbances and
unrepresentative dilution of the gas stream.
3. Turn on the pump and immediately adjust the sample flow to attain isokinetic
conditions. Nomographs, calculator programs, and routines are available to
aid in the rapid determination of the orifice pressure drop corresponding to
the isokinetic sampling rate. If the nomograph is designed as shown in APTD-
0576 it can be used only with an Type S pitot tube which has a Cp coefficient
of 0.85 ± 0.02 and when the stack gas dry molecular weight (Ms) is 29 i 4.
If Cp and Ms are outside these ranges, do not use the nomograph without
o
X""X
f J
^—
-------
Section No. 3.19.4
Date September 3, 1992
Page 7
compensating for the differences. Recalibrate isokinetic rate or reset
nomograph if the absolute stack temperature (Ts) changes more than 10%.
4. Take other readings required by Figure 4.2 at least once at each sampling
point during each time increment.
5. Record the dry gas meter readings at the end of each time increment.
6. Repeat steps 3 through 5 for each sampling point.
7. Turn off the pump, remove the probe from the stack, and record the final
readings after each traverse.
8. Conduct the mandatory posttest leak check (Subsection 4.2.5) at the
conclusion of the last traverse (after allowing the nozzle to cool). Record
any leakage rate. Also, leak check the pitot lines (Method 2, Section 2.1);
the lines must pass this leak-check to validate the velocity pressure data.
9. Disconnect the probe, and then cap the nozzle and the end of the probe with
polyethylene or equivalent caps.
-------
Plant
City
Location
Operator
Date
Run No.
Stack dia. mm (in).
Sample box No.
Meter box No.
Meter AH@
Meter calibration (Y)
Pitot tube (Cp)
Probe length
Probe liner material
Probe heater setting
Ambient temperature _
Barometric press (Pb)
Assumed moisture
Static press. (Pa)
C Factor
Reference AH@
Sheet
Nozzle ID No.
of
ft Nozzle diameter
Thermometer No.
Final leak rate
°F Vacuum during leak-check
iron (in)
m3/min (cfm)
mm (in) Hg
mm (in) Hg Filter No(s).
%H2O Remarks:
mm (in) Hg
traverse
point
number
Sampling
time,
(6) , min
\
[_..
Total
Clock
time,
(24 h)
._
Vacuum,
mm
(in) Hg
Max
Stack
temp
(Ts)
°C (°F)
•
Avg
Velocity
head
(APs)
mm
(in) Hg
Press
across
orifice
meter
(AH) , mm
(in) Hg
Gas sample
volume (Vm) ,
m3 (ft3)
"J
--
(Total
i
Dry gas meter
temperature
Inlet | Outlet
°C(°F) °C(°F)
Avg | Avg
Gas temp
leaving
impinger
°C(°F)
Max
Filter
temp
°C(°F)
0> D» rr
1 CD > o
..._. *rt a
H
L. ^M
OJ «>
L-
ID
to
Figure 4.2. Method 101A field sample data form.
o
o
o
-------
'Section 3.19.4
Date April 3, 1992
Page 9 -i
During the test run, a sampling rate of 10% of the isokinetic rate must be
maintained unless otherwise specified by the Administrator. The sampling rate must be
adjusted at any sampling point if a 20% variation in velocity pressure occurs.
Periodically during the test, observe the connecting glassware—from the probe,
through the filter, to the first impinger—for water condensation. If any is evident,
adjust the probe and/or filter heater setting upward until the condensation is
eliminated; add ice around the impingers to maintain the silica gel exit temperature
at 20 °C (68 °F). : -
The manometer level and ; zero should also be checked periodically during each
traverse. Vibrations and temperature fluctuations can cause the manometer zero to
Shift. " -> ' -c- '••• .....--:. . ,:
4 .4 Sample Recovery
The reference method (Section 3.19.10) requires that the sample be recovered from
the probe, from all glassware preceding the filter, from the front half of the filter
holder, from the filter, and from the impingers and connecting glassware in an area
sheltered from wind and dust to prevent contamination of the sample. Begin proper
cleanup procedure as soon as the probe is removed from the stack at the end of the
sampling period. Allow the probe to cool. When-art can be safely handled, wipe off any
external particulate matter near the tip of the probe nozzle, and place a cap over it.
Do not cap off the probe tip tightly while the sampling train is cooling because the
resultant vacuum could draw liquid out from the impingers. Before moving the sample
train to the cleanup site, remove the probe from the train, wipe off the silicone
grease, and cap the open outlet of the probe and the inlet of the sample train.
Be careful not to lose any condensate that might be present. Wipe off the
silicone grease from the impinger. Use either ground-glass stoppers, plastic caps, or
serum caps to close these openings. The capped-off impinger box .and the capped
sampling probe can be transported to the cleanup area without risk of losing or
contaminating the sample. Transfer the probe, impinger assembly, and (if applicable)
filter assembly to a cleanup area that is clean, protected from the wind, and free of
Hg contamination. The ambient air in laboratories located in the immediate vicinity
of Hg-using facilities is not normally free of Hg contamination. Inspect the train
before and during disassembly, and note any abnormal conditions. .->,.,;••._.
Precautions It is extremely important that all sample recovery personnel wear
safety glasses and gloves due to the dangers associated with impinger solutions and
recovery solutions. ..-".-••!•
The following sample recovery sequence includes (1) recovery-of the sample from
the impingers using KMnO<, Container 1; (2) recovery of any residual brown deposits
from the impingers using water, Container 1; (3) recovery of the sample from the probe
and connecting glassware using KMnO«, Container No.l; (4) recovery of any residual
brown deposits from the probe and connecting glassware using water, Container No.l; (5)
recovery of any residual brown deposits from sample train components not removed by
water with HC1, Container 1A; recovery of silica gel. Container 2; (6) recovery of the
filter, Container No. 3; (7) collecting a. filter blank, Container No.,4,; (8) collecting
an KMnO< reagent blank, Container No. 5; (9) collecting a water reagent blank, and (10)
collecting a HC1 reagent blank.
4.4.1 Impinger Contents (Container Nos. 1 and 1A)—Recover the samples follows: c
1. Note the color of the reagent in each of the impingers and record the color
on the Sample Recovery Data Form. If the color of the KMnO« in the last
impinger has changed from the purple color, the sample run will be considered
invalid and must be repeated. •• If all the impinger solution has been
-------
o
. Section 3.19.4 .--.-<
-•• " Date April 3, 1992
Page 10
oxidized, the tester should (1) reduce the sample time or volume if the
reduced time or volume will comply with the applicable regulations, (2) add
another impinger containing KMnO<, or (3) use two sample trains per sample
run.
2. Using a properly cleaned graduated cylinder, measure the liquid in the first
three impingers to within 1 ml. Record the volume of liquid on the Sample
Recovery and Integrity Data Form. This information is needed to calculate
the moisture content of the effluent gas. (Use only graduated cylinders and
glass storage bottles 'that have been precleaned as in Section 3.19.3.)
3. Place the contents of the first three impingers in a properly cleaned,
1000-ml glass sample bottle (Container No. 1). Record the data on the sample
recovery data form.
4. Prior to recovering the sample, place 400-ml of fresh KMnO« in a graduated
cylinder for sample recovery. This solution is used to recover sample from
the probe nozzle, probe fitting, probe liner, and front half of the filter
holder (if applicable) and impingers (sample-exposed surfaces). Rinse the
impingers with a portion (about 100 ml) of the 400 ml of fresh 4% KMnO<
solution to assure removal of all loose particulate matter from the
impingers; add all washings to the 1000-ml glass sample bottle (Container No.
1) .
5. To remove any residual brown deposits; on the glassware following the
permanganate rinse, carefully rinse all the sample-exposed glassware with
approximately 100 ml of water. Add this rinse to Container No. 1. The
impingers should only require about 50 ml of the 100 ml of water.
6. If no visible deposits remain after this water rinse, do not rinse with 8 N
HC1. However, if deposits remain on the glassware after the water rinse,
place 25 ml of 8 N HC1 in a graduated cylinder. Wash impinger walls and
stems with this 25 ml of 8 N HCl as follows: Place 150 ml of water in a
sample container labeled Container No. 1A. Use only a total of 25 ml of 8 N
HCl to rinse all impingers. Wash the impinger walls and stem with the HCl by
turning and shaking the impinger so that the HCl contacts all inside
surfaces. Pour the HCl wash carefully while stirring into Container No. 1A.
Rinse all glassware that was exposed to HCl with 50 ml water, and add water
rinse to Container No. 1A. Label the sample bottle and record the sample
number on the Sample Recovery Data Form. The separate container is used for
safety reasons.
4.4.2 Probe and Connecting Glassware (Container No. Ij-The same sample bottle
(Container No. 1) as used above for the impinger contents and sample rinse is usually
adequate for the collection of all the rinses. Recover the sample from the probe liner
and connecting glassware as follows:
1. Clean the outside of the probe, the pitot tube, and the nozzle to prevent
particulates from being brushed into the sample bottle. Take care that dust
on the outside of the probe or other exterior surfaces does not get into the
sample during the quantitative recovery of the Hg (and any condensate) from
the probe nozzle, probe fitting, probe liner, and front half of the filter
holder (if applicable).
2. Carefully remove the probe nozzle and rinse the inside surface (using a ny
bristle brush and several KMnO< rinses) into the sample bottle (Container1- No f
1) .
3. Clean the compression fitting by the same procedure. Rinse all sampleexposed
glassware components with the total of 400 ml of fresh 4% KMnO< solution as
O
-------
Section 3.19.4
Date April 3, 1992
Page 11
measured above. Add these washings to the 1000-ml glass sample bottle
(Container No. 1).
4. After the KMnO4 rinse, use a small portion of the remaining 100 ml of water
to rinse the nozzle and connecting glass after the KMnO4 rinse. Add the
rinses to Container 1.
The following probe rinsing procedure should be performed by two people to
preclude sample loss. The rinsing procedures for the probe liner and connecting
glassware is as follows:
1. Rinse the probe liner by tilting and rotating the probe while squirting fresh
4% KMnO4 solution into the upper (or nozzle) end to assure complete wetting
of the inside surface.
2. Allow the KMnO, solution to drain into the sample bottle (Container 1) using
a funnel to prevent spillage.
3. Hold the probe in an inclined position and squirt KMnO4 solution into the
upper end while pushing the probe brush through the liner with a twisting
motion, and catch the drainage in the sample bottle. Repeat the brushing
procedure three or more times until a visual inspection of the liner reveals
no particulate remaining inside.
4. Rinse the liner once more with KMn04 solution.
5. Rinse the brush with KMn04 solution into^Container 1 to remove all sample that
is retained by the bristles. >
6. Rinse the probe liner with the remaining 100 ml of water into Container 1.
7. Wipe all the connecting joints clean of silicone grease, and clean the inside
of the front half of the filter holder by rubbing the surface with a nylon
bristle brush and rinsing it with KMn04. Repeat the procedure at least three
times or until no particles are evident in the rinse.
8. Make a final rinse of the filter holder and brush.
9. Clean any connecting glassware which precedes the filter holder, using Steps
5 and 6.
After all washings have been collected in Container No. 1, tighten the lid on the
container to prevent leakage during shipment to the laboratory. It is recommended that
the lid have a No. 70-72 hole drilled in the container cap and Teflon liners for
pressure relief. Mark the height of the fluid level to determine whether leakage
occurs during transport. Label the container to identify its contents clearly, and
note it on the Sample Recovery Data Form.
4.4.3 Silica Gel (Container No. 2)—Note the color of the indicating silica gel to
determine whether it has been completely spent, and make a notation of its condition
on the sample recovery data form, Figure 4.3.
1. Transfer the silica gel from the fourth impinger to its original container
using a funnel and a rubber policeman, and seal the container. It is not
necessary to remove the small amount of dust particles that may adhere to
the impinger wall; since the weight gain is used for moisture calcula-
tions, do not use water or other liquids to transfer the silica gel.
2. Determine the final weight gain to the nearest 0.5 g, if a balance is
available.
4.4.4 Filter ("Container No. 3)-Carefully remove the filter (if used) from the filter
holder, place it in a 150-ml glass sample bottle, and add 20 to 40 ml of 4% KMnO^ to
merge the filter. If it is necessary to fold the filter, be sure that the
'articulate cake is inside the fold. Carefully transfer, to the 150-ml sample bottle,
any particulate matter and filter fibers that adhere to the filter holder gasket by
using a dry Nylon bristle brush and a sharp-e|dged blade. Seal the container. Clearly
-------
Section 3.19.4
Date April 3, 1992
Page 12
label the container to identify its contents. Mark the height of the fluid level to
determine whether leakage occurs during transport.
4.4.5 Filter Blank (Container No. 4)—If. a filter is used for testing, initially take
an unused filter for each field test series and label as a filter blank. Treat the
filter blank in the same manner as described in Subsection 4.3.4 above.
4.4.6 Absorbing Solution Blank (Container No. 5;-For a blank, place 650 ml of 4% KMn04
absorbing solution in a 1000-ml sample bottle. If the 100 ml water rinse was used
during recovery, carefully add a second 100 ml portion of water to Container No. 5.
It is recommended that the lid have a No. 70-72 hole drilled in the container cap and
Teflon liners for pressure relief. Mark the height of the fluid level to determine
whether leakage occurs during transport. Label the container as the KMnO4 blank, and
seal the container.
4.4.7 8 N HC1 Blank (Container No. 6)-It 8 N HCl was used (Container 1A) to remove any
residual brown deposits remaining after rinsing sample-exposed glassware with fresh 4%
KMnO, solution and water, place 25 ml of 8 N HCl used for removing the deposits in a
separate sample container (Container No. 6) containing 200 ml of water. Mark container
as the HCl blank, and seal the container. -f~
\
4.5 Sample Logistics and Packing Equipment
Follow the sampling and sample recovery procedures until the required number of
runs are completed and blank samples are labeled. Log all data on the Sample Recovery
and Integrity Data Form, Figure 4.1. At the conclusion of the test:
1. Check all rinses and filters for proper labeling (time, date, location, test
run number, and any other pertinent documentation) . Be sure that blanks have
been set aside and labeled.
2. If possible, make a copy of the field data form(s) in case the originals are
lost.
3. Examine all sample containers for damage and ensure that they are properly
sealed for transport to the base laboratory. Ensure that the containers are
labeled properly for shipping to prevent loss of samples or equipment.
4. Review the field sampling data form and any other completed data forms to
ensure that all data have been recorded and that all forms are present.
4.6 Systems Audit
A Method 101A sampling and sample recovery checklist is presented in Figure 4.3.
o
o
o
-------
Section 3.19.4
Date April 3, .1992-
Page 13
Date Time • Operator Observer
Method 101A Sampling Procedures
Probe Nozzle: stainless steel glass
Button-hook elbow size
Cleaned according to sampling protocol? __
Sealed with Teflon tape or other cover?
Probe liner: borosilicate quartz _________ other
Cleaned according to sampling protocol? "
Openings sealed with Teflon tape?
Probe heating system:
Checked? Temperature Stable?
Pitot tube: Type S Other
Properly attached to probe (no interference to nozzle)?
Modi f i cat ions: ;
Pitot tube coefficient \
ijpjpjjj/ Differential Pronsure Gauge: Inclined manometers
Magnahelics Ranges
Other Ranges
Cyclone (inlet only): borosilicate glass other
Cleaned according to sampling protocol?
Filter Holder: borosilicate glass other
Frit material: glass Teflon other
Gasket material: silicone other
Cleaned according to sampling..protocol?
Sealed with Teflon tape or glass caps?
Filter type (a):
Cleaned according to sampling protocol?
Impinger Train: number of impingers
Cleaned according to sampling protocol?
Contents: 1st 2nd 3rd
4th 5th 6th
Impinger weights recorded? ;
Proper connections? —
Modifications
Silica gel: type new? used?
Figure 4.3. Field observation of Method 101A sampling and recovery.
-------
Section 3.19.4
Date April 3, 1992
Page 14
o
Date Time Operator Observer
Method 101A Sampling Proceduren
Procedure
Barometer: mercury aneroid other
Can Doncity Determination: temperature sensor
pressure gauge
Temperature sensor properly attached to probe?
Recent Calibrations: pitot tubes
meter box thermocouples/thermometers
Filters checked visually for irregularities?
Filters properly centered? labelled?
Sampling site properly selected?
Nozzle size properly selected?
Proper sampling time selected or calculated?
All openings of sampling train sealed (pretest
and posttest)?
Impingers, filter holder, probe, and nozzle assembled?
Cyclone attached (inlet only)?
Pitot lines checked for leaks and plugging?
Meter box leveled? Manometers zeroed?
AH@ from most recent calibration
Nomograph setup correctly? K factor
Pretest leak-check conducted? __Leakage rate?
Care taken to avoid scraping nipple or stack wall?
Effective seal around probe when in-stack?
Probe moved to traverse points at proper time?
Figure 4.3. (Contined)
O
o
-------
Section 3.19.4
Date April 3, 1992
Page 15
Date
Time
Operator
Observer
Method 101A Sampling Proccduroo
Nozzle and pitot tubes kept parallel to stack at all times?
Filter(s) changed during run?
Any particulate lost during filter change?
Data forms completed and data recorded properly?
Nomograph setting changed with significant change in the stack temperature?
Velocity pressure and orifice pressure recorded.accurately?
—r**
Leakage rate
Posttest leak-check conducted?
at inches of mercury
Orsat analysis?
Stack
Integrated
Approximate stack temperature
Percent isokinetic calculated
Comments
Gas sample volume
Figure 4.3. (Continued)
-------
Section 3 .19.4
Date April 3, 1992
Page 16
Date Time Operator Observer
Mothod 101A Sample Rocovory
Reagentc:
Bruohoo: Teflon bristle other
Cleaned according to sampling protocol?
Waoh bottloo: glass other
Cleaned according to sampling protocol?
Storage containoro: glass? other?
Cleaned according to sampling protocol?
Teflon cap liner? Leak free?
Small hole in cap to relieve pressure?
Filter containorc: borosilicate glass other
Cleaned according to sampling protocol? -~
Graduated cylinder: borosilicate glass other
Subdivisions of graduated cylinder <2 ml?
Cleaned according to sampling protocol?
Balance typo: Calibrated?
Probe allowed to cool sufficiently?
Probe and sample train openings covered?
Clean-up area(s) used
KMnO4 Volume: Was 400 mL of KMnO< measured for recovery?
Filter handling: tweezers used? surgical gloves?
Any particulate lost?
KMnO< added to filter?
Probe handling: KMn04 rinses Brushed?
H,O rinses Brushed?
Rocovory of probo: probe nozzle probe fitting
probe liner front half of filter holder _
Figure 4.3. (Continued)
o
o
o
-------
Section 3 .19.4
Date April 3, 1992
Page 17
Date Time Operator . Observer
Method 101A Sample Recovery (cont)
HC1 Volume: Was 25 mL of HC1 measured for recovery?
Xmpingor handling: weighed? volumed?
KMn04 rinses H2O rinse
HC1 rinses
Blankn collected: filter
KMnO< (650 mL)
HC1 (25 mL in 200 mL of H2O)
Container No. 1: Sample No. 400 mL KMnO4 rinse.
Impinger contents Impinger Rinse
Probe rinse Nozzle rinse ^Z
Container No. 1A: Sample No. 25 mL HCl
Impinger rinse
11 'v
ontainer No. 2 Silica gel: color? condition? weighed?
Samples labeled and stored properly?
Liquid levels marked?
Remarks: __
Figure 4.3. (Continued)
-------
Section 3.19.4
Date April 3, 1992
Page 18
o
TABLE 4.1. ACTIVITY MATRIX FOR ON-SITE MEASUREMENT CHECKS
Apparatus
Acceptance limits
Frequency and method
of measurement
Action if
requirements
are not met
Preliminary
determinations and
measurements
Determine the
moisture content of
stack gas
Determine flow rate
of stack gas
Determine stack
temperature
Determine stack
dimensions
Determine dry
molecular weight of
stack gas
Select sampling time
S minimum total sam-
pling time in
applicable emission
standard; number of
minutes between read-
ings should be an
integer
Once each field test;
use wet bulb/dry bulb
thermometer,'Method 4,
or sling psychrometer
Once each field test,
using Method 1
Prior to and during
sampling
Prior~to sampling,
using tape measure
Once each field test,
Method 2; if inte-
grated gas sample is
required, Method 3
Prior to sampling
Complete
Complete
Complete
Complete
Complete
O
Complete
Preparation of
collection train
Assemble train
according to
specifications in
Figure 1.1 and Sec.
3.18.4 Subsec. 4.3.3
Leak-check; Leak
rate < 4% or 0.00057
mVmin (0.02 ft3/
min) , whichever is
less
Before each sampling
run
Complete
Leak-check before sam-
pling by plugging the
nozzle or inlet to
first impinger and by
pulling a vacuum of
380 mm (15 in) Hg
Correct the leak
O
(Continued)
-------
Section No. 3.19.5
' Date September 3, I9B2
Page 1
5.0 POSTSAMPLING OPERATIONS
The postsampling operations include postsampling calibration checks of
sampling equipment and analysis by atomic absorption spectrophotometry techniques. The
sample analysis includes calibrations and performance checks. Checklists for
monitoring the postsampling operations are provided at the end of this section. Table
5.1 at the end of this section summarizes the QA activities associated with the
postsampling operations.
5.1 Calibration Checks of Sampling Equipment
Posttest checks will have to be made on most of the sampling apparatus. These
checks will include three calibration runs at a single orifice meter setting, cleaning,
and/or routine maintenance. Cleaning and maintenance are discussed in Section 3.19.7
and in APTD 0576. Figure 5.1 can be used to record the posttest checks.
5.1.1 Metering System—The metering system has two components that must b=
checked-the dry-gas meter and the dry-gas meter thermometer(s).
The dry-gas meter thermometer(s) should be compared with the ASTM mer-
cury-in-glass thermometer at room temperature. If the two readings agree within 6 CC
(10.8 °F) , they are acceptable; if not, the thermometer must be recalibrated according
to Subsection 2.2 of Section 3.19.2 after the posttest check of the dry-gas meter. Fcr
calculations, use the dry-gas meter thermometer readings (field or recalibraticr.
values) that would give the higher temperatures.':'That is, if the field readings are
higher, no correction is necessary, but' if the rec'alibra'tion value is higher, add the
difference .in the two readings to the average dry-gas meter temperature reading.
The posttest check of the dry-gas meter is described in Section 3.19.2. The
metering system should not have any leaks that were corrected prior to the posttesr
check. If the dry-gas meter calibration factor (Y) deviates by <5% from the initial
calibration factor, the dry-gas meter volumes obtained during the test series are
acceptable. If Y deviates by >5%, recalibrate the metering system (Section 3.19.2;.
For the calculations, use the calibration factor (initial or recalibration) that yields
the lower gas volume for each test run.
5.1.2 Stack Temperature Sensors—The stack temperature sensor readings should be
compared with the reference thermometer readings.
For thermocouple (s), compare the thermocouple and reference thermometer values
at ambient temperature. If the values agree within 1.5%!of the absolute temperature,
the calibration is considered valid. If the values do not agree within 1.5%, recali-
brate the thermocouple as described in Section 3.19.2 to determine the difference (A75
at the average stack temperature (T.) . NOTE: This comparison may be done in the field
immediately following the tests.
For thermometers, compare the reference thermometer:
1. At ambient temperatures for average stack temperature below 100 °C (212
-------
Section No. 3.19 .5
Date September 3, 1992 /"""N
Page 2 I )
Plant Calibrated by
Meter box number Date
Dry-Gas Meter
Pretest calibration factor, Y .. .. (within 2%)
Posttest check, Y* (within 5% of pretest value)
Recalibration required? ' yes ____________________ no
If yes, calibration factor, Y (within 2%)
Lower calibration factor, Y for calculations (pretest or posttest)
Dry-Gas__ Meter Thermometers
Was a pretest temperature correction used? yes _ no
If yes, temperature correction (within 5.4 °F over range)
Posttest comparison with mercury-in-glass thermometer? * (within 10.8 °F at ambient
temperature) , °F
Recalibration required? yes _no
Recalibration temperature correction? ' (within 5.4 °F over range)
If. yes, no correction necessary for calibration if meter thermometer temperature
is higher, if calibration temperature is higher, add correction to average meter
temperature for calculations.
Stack Temperature Sensor
o
Was a pretest temperature correction used? yes no
If yes, temperature correction °F (within 1.5% in. °R over range)
Average stack temperature of compliance test, T, °R .
Temperature of reference thermometer or solution °R (within 10% of T.)
Temperature of stack temperature for recalibration °R
Difference between reference and stack thermometers, AT, °R
Do values agree within 1.5%?* yes no
If yes, no correction necessary for calculations.
If no, calculation must be done twice—once with the recorded values and once with
the average stack temperature corrected to correspond to the reference
temperature differential (AT.) . Both final results must be reported.
Barometer
Was the pretest field barometer correct? yes no
Posttest comparison?* in. Hg (within 0.1 in. Hg)
Was recalibration required? yes no
If yes, no correction necessary for calculations when the field barometer has a
lower readings; if the mercury-in-glass reading is lower, subtract the difference
from the field data readings for the calculations.
Figure 5.1. Posttest calibration checks.
O
-------
: Section No. 3.19.5
Date September 3, 1992
Page 3
2. In boiling water for stack temperatures from 100 °c to 200 °C.
3. In a boiling liquid with the boiling point above 200 °C for stack
temperatures between 200 to 405 °C. For stack temperatures above 405
°C, compare the stack thermometer with a thermocouple at a temperature
within 10% of the average stack temperature. If the absolute values
agree within 1.5%, the calibration is considered valid. If not,
determine the error (AT.) to correct the average stack temperature.
5.1.3 Barometer—The field barometer should be compared to a Hg-in-glass barometer.
If the readings agree within 5 mm (0.2 in.) Hg, the field readings are acceptable; if
not, use the lesser calibration value for the calculations. If the field barometer
reads lower than the Hg-inglass barometer, the field data are acceptable. If the
Hg-in-glass barometer gives the lower reading, use the difference in the two readings
(the adjusted barometric value) in the calculations.
5.2
Sample Preparation
Field samples and reagent blanks should be'prepared concurrently, if possible.
Check the liquid level in each container to see whether liquid was lost during
transport. If a noticeable amount of leakage occurred, either void the sample or use
methods subject to the approval of the Administrator-to account for the losses. Record
the findings of the liquid level check on the sample1 preparation data form, Figure 5.2,
or another suitable form. Then follow the procedures below.
5.2.1 Containers Wo. 3 and
following procedures apply:
No. 4 (Filter and Filter Blank)—1£ a. filter; is used, the
Place the contents, including the filter, of Container No. 3 in a
separate, properly cleaned, and uniquely identified 250-mL beaker.
Using three rinses of approximately 10 mL of water, complete the sample
transfer from the container. Record the beaker number with the run
number on the sample preparation data form.
Place the contents of Container No. 4 in a properly cleaned 250-mL
beaker. Label it as the sample filter blank or as another suitable
name. Use three rinses of approximately 10 mL of water for the sample
transfer. Record the name on the sample preparation data form.
Heat the beakers in a laboratory hood on a steam bath until most of the
liquid has evaporated. Do not take to dryness. Do not use direct
heating on a hot plate. Record the completion of the step on the sample
preparation data form.
Add 20 mL of concentrated HN03 to each beaker, cover each beaker with a
watch glass, and heat on a hot plate at 70 °C • for 2 h in a laboratory
hood. Record completion of this step on the sample preparation data
form.
-------
Section No. 3.19.5
Date September 3, 1992
Page 4
Sample Preparation Data Form
Date _ Plant Name _ Sampling Location
Sample Preparation Checks
Sample Integrity Check: Have containers leaked?
Container 1 _ 4 _
1A _ 5 _
2 _ 6 _
3
Run Run Run Blank
123
Place a check to indicate completion
of step or record data as indicated.
Preparation of Filter Digest: Container No. 3
Sample No. for each 250-mL beaker
Contents added to a glass 250-mL beaker?
Heated carefully to near dryness
(not dryness) using a steam bath?
Volume of HNO3 added to beaker 25 mL?
Covered with watch glass?
Heated at 70 °C on hot plate for 2 h?
How was temperature monitored?
Filtered through Whatman 40 paper ?
Date
Time
Rinsed beaker residue carefully through
the filter?
Saved filtrate?
Preparation of Sample No. A.I:
Are Container No. 1 contents <1000 mL?
If so, volume, mL
Are Container No. 1 contents filtered through
Whatman 40 paper?
Filter saved?
Filtrate added to mL glass volumetric flask?
Filter digest (above) added to flask with
Container No. 1 filtrate?
Completion of Sample No. A.I?
Date
Time
o
o
Figure 5.2. Sample preparation data form.
-------
Section No. 3.19.5
Date September 3, 1992
Page 5
Run Run Run Blank
1 • 2 ••..-. 3
Preparation of Sample No. HCl A.2:
25 mL of 8N HC1 added to filter saved from
preparation of Sample No. A.I?
How was HC1 added?
Digestion started, Time
Date
Digestion ^completed, Time
Date
HC1 digest dilution volume, mL
Preparation of Filter Blank;
Container 4 contents added to 250-mL beaker
Heated carefully to near dryness
(not dryness) using a steam bath? -^"n-
Volume of HN03 added to beaker 25 mL?
Covered with watch glass?
Heated at 70 °C on hot plate for 2 h?
How was temperature monitored?
Preparation of Sample A.I Blank;
Are Container No. 5 contents diluted to same '
volume as Container No. 1 contents?
Filtered through Whatman 40 paper?
Filter saved?
Filtrate added to 1000-mL glass
volumetric flask?
Filter blank (Container No. 4) digest (above)
added to same volumetric flask?
Time of completion of Sample No. A.I
Preparation of Sample No. HCl A.2 Blank;
25 mL of 8N HCl added to filter saved from
preparation of Sample No. A.I blank?
How was HCl added?
Time 24-h digestion started?
Date
Time
Time 24-h digestion completed?
Date
Time
HCl digest was diluted to 500 mL using glass
volumetric flask?
Figure 5.2. (Continued)
-------
Section No. 3.19 .5
Date September 3, 199
Page 6
Note: The analysts should use gloves and safety glasses and should avoid skin
contact and breathing the fumes from the HN03.
5. Remove the beaker from the hot plate and filter the solution from the
digestion of the contents of Container No. 3 through a separate Whatman
40 filter paper into a properly cleaned and identified (the same sample
identification number can be used) sample container using a vacuum
filtering system. Use three rinses of approximately 10 mL of water for
the sample transfer. The filtration should be conducted in a laboratory
• hood. Record the completion of this step on the sample recovery data
form.
6. Save the filtrate for addition to the Container No. 1 filtrate, as de-
scribed in Subsection 5.2.2. Discard the filter.
7. Filter the solution from the digestion of the contents of Container No.
4 (sample filter blank) through Whatman 40 filter paper, as described
above in Step 5, and save the filtrate for addition to the Container No.
5 filtrate, as described in Section 5.2.2. Discard the filter.
5.2.2 Container No. 1 (Impingers, Probe, and Filter Holder) and. If Applicable, 1A
(HCl Rinse)—The KMnO4 impinger solution and rinse and HCl rinse (if applicable) are
prepared as follows: -*~
Note: Because of the hazardous nature of the HNO3 and HCl solutions, analysts
must wear gloves and safety glasses and should avoid skin contact and breathing they
fumes from HN03 and HCl. The HNCX, and KMnO4 solutions should not come in contact wit*
oxidizable matter.
KMnO, Impinqer Solution and Sample Recovery Rinse
1. To remove the brown MnO2 precipitate, filter the contents of Container
No. 1 through a Whatman 40 filter into a properly cleaned and identified
1-L volumetric flask. Use three rinses of approximately 10 mL of water
for the sample transfer.
2. Save the filter for digestion of the brown MnO2 precipitate, as
described in steps 6 through 9 below, and record the date and time the
filtration step was completed on the sample preparation data form.
3. Add the sample filtrate from Container No. 3 produced in Subsection
5.2.1 above to the appropriate 1-L volumetric flask from Step 1, and
dilute to volume with water. If the combined filtrates are greater than
1000 mL, determine the volume to the nearest mL and record the volume
on the sample preparation data form. This volume will be used to make
the appropriate corrections for blank subtractions and emissions
calculations.
4. Mix thoroughly. The filtrate will be referred to as Analysis Sample No.
A.I.
5. The Analysis Sample No. A.I must be analyzed for Hg within 48 h after
completion of the filtration step. If the sample is not analyzed within
this period, steps 1 through 4 must be repeated, the additional Whatman
40 filter paper will be digested as described below in steps 6 through
9, and the digestion will be added to the sample.
Whatman 40 Filter and MnCs Precipitate
6. Place the saved filter, which was used to remove the brown MnQf ^
precipitate, into a container of appropriate size. Submerge the filte\>_y'
with 25 mL of 8 N HCl and allow it and the brown residue to digest for
a minimum of 24 h at room temperature. Record the date and time for the
beginning of the digestion on the sample preparation data form.
-------
Section No. 3.19.5
Date September 3, 1952
Page 7
Whatman 40 Filter. MnO, Precipitate, and HC1 Rinse
7. Filter the contents of Container No. 1A, HC1 rinse (if applicable)
through a Whatman 40 filter into "a properly cleaned and identified
500-mL volumetric flask. Use three rinses of approximately 10 mL of
water for the sample transfer. Record completion on the sample
preparation data form.
8. Filter the digestion of the brown MnO2 precipitate and Whatman filter
from Step 6 into the 500-mL volumetric flask from Step 7. Use three
rinses of approximately 10 mL of water for the sample transfer. Record
the date and time of the filtration on the sample preparation data fora.
9. Dilute to volume with water. This solution will be referred to as
Analysis Sample No. HC1 A.2. Save the solution for Kg analysis as
described in Subsection 5.3.4 below. Discard the filters.
5.2.3 Containers No. 5 (Absorbing Solution Blank) and No. 6 (HC1 rinse blank)-Tne
procedures for preparing the blank solutions are described below:
Note: The same precautions should be taken with the blank solutions as were taken with
the sample solutions. The sample blanks have been designed to allow easy blank
subtraction from the sample. The volume of all solutions and the number of filters are
identical to the field samples. Therefore, the blank sample must be prepared at the
same time and in the same manner as the field samples.
KMnO4 Reagent Blank Solution and Sample Recovery Blank Rinse
1. Treat Container No. 5 (650 mL of blank absorbing solution) the same as
Container No. 1 (described in steps 1 through 5 in Subsection 5.2.2).
Filter Blank
2. Add the filter blank filtrate from Container No. 4 (completed in steps
1 through 7 of Subsection 5.2.1 above) to the 1-L volumetric flask
(containing Container No. 5 filtrate), and dilute to volume. Mix
thoroughly.
3. This solution will be referred to as Analysis Sample No. A.I blank.
4. Analysis Sample No. A.I blank must be analyzed for Hg within 48 h after
the completion of the filtration step.
Whatman 40 Filter and KMnO4 Reagent Blank Precipitate
5. Digest any brown precipitate remaining on the filter from the filtration
of Container No. 5 by the same procedure described in step 6 ir.
Subsection 5.2.2 above.
Whatman 40 Filter, KMnO4 Blank Precipitate, and Blank HCl Rinse
6. Filter the contents of Container No. 6 by the same procedure described
in steps 7, 8, and 9 in Subsection 5.2.2 and combine into the 500-nl
volumetric flask with the filtrate from the digested KMnO< blar.k
precipitate. The resulting 500-mL combined dilute solution will be
referred to as Analysis Sample No. HCl A.2 blank. NOTE: As discussed ir.
Subsection 5.3.4 below, when analyzing samples A.I blank and HCl A.2
blank, always begin with 10-mL aliquots; this note applies specifically
to blank samples.
5.3
Analysis
Precise and accurate analysis requires that the Hg analysis system bs
calibrated properly, which includes preparing calibration standards and field samples.
-------
Section No. 3 .19.5
Date September 3, 1992
Page 8
For Method 101A, spectrophotometer calibration is conducted in conjunction with
analyzing the field samples (and QA samples). This section presents the steps for
analyzing the field samples and includes preparing sample and field blanks, as well as
describing how to quality control (QC) samples. It discusses the relationship between
analyzing the field samples and preparing the calibration curve.
5.3.1 Instrument Setup-Before use, clean all glassware, both new and used, as
follows: brush with soap and tap water, liberally rinse with tap water, soak for 1 h
in 50% HN03, and then rinse with deionized distilled water.
Flow Calibration—Assemble the aeration system as shown in
Figure 5.3. Set the outlet pressure on the aeration gas cylinder regulator to a
minimum pressure of 500 mm Hg (10 parts per square inch [psi]); use the flow metering
valve and a bubble flow meter or wet-test meter to obtain a flow rate of 1.5 ± 0.1
L/min through the aeration cell. After the flow calibration is completed, remove the
bubble flow meter from the system.
Optical Cell Heating System Calibration-Using a 25-mL graduated cylinder, add
25 mL of water to the bottle section of the aeration cell. Attach the bottle section
to the bubbler section of the cell. Connect the fferation cell to the optical cell and,
while aerating at 1.5 L/min, determine the minimum variable transformer setting
necessary to prevent condensation in the optical cell and in the connecting tubing.
(This setting should not exceed 20 volts.)
Wavelength Adjustment—Set the spectrophotoraeter wavelength at 253.7 nm and
make certain that the optical cell is at the minimum temperature needed to prevent
water condensation.
Recorder Adjustment—The Hg response may be measured by either peak height or
peak area. Peak height determinations may be performed manually by counting the
recorder paper divisions for a given peak from a best-drawn baseline. The peak height
from the baseline also may be measured conveniently using a millimeter ruler. Peak
area measurements are most conveniently accomplished electronically using an integrator
or similar device. For peak height determinations, set the recorder scale as follows:
Note: The temperature of the solution affects the rate at which elemental Hg is
released from a solution and, consequently, it affects the shape of the generated peak
as well as the peak height. Therefore, to obtain reproducible results using peak
height, bring all solutions to room temperature before use.
o
O
o
-------
Section No. 3.19.5
Date September 3, 1992
Page 9
NEEDLE VALVE FOR
FLOW CONTROL
CYLINDER
CELL
FLOW
METER
EXIT ARM
Hfr
TO HOOD
TO VARIABLE TRANSFORMER
PTICAL CELL
MAGNETIC STIRRING BAR
MAGNETIC STIRRER
Figure 5.3. Schematic of aeration system.
-------
Section No. 3 .19 .5
Date September 3, 195
Page 10
1. Place a Teflon-coated stirring bar in the bottle. Using a 25-cL,
graduated cylinder, add 25 mL of laboratory pure water to the aeration
cell bottle. Pipet 5.0 mL of the working Hg standard solution to the
aeration cell.
2. Add 5 mL of the 4% KMn04 absorbing solution followed by 5 mL of 15% HNO:.
and 5 mL of 5% KMn04 to the aeration cell and mix well using a swirling
motion.
3. Attach the bottle to the aerator making sure that: (1) the exit am
stopcock is closed, and (2) there is no aeration gas flowing through the
bubbler.
4. Through the side arm, add 5 mL of sodium chloride hydroxylamine solution
in 1-mL increments until the solution is colorless.
5. Through the side arm, add 5 mL of the Tin (II) reducing agent to the
aeration cell bottle and immediately stopper the side arm.
6. Stir the solution for 15 s and turn on the recorder or integrator.
7. Open the aeration cell exit arm stopcock and initiate the gas flow.
8. Determine the maximum height (absorbance) of the standard and set this
value to read 90% of the recorder full scale.
5.3.2 Analytical Calibration Curve-After setting the recorder scale (Section 5.3.1! ,
the calibration is performed. To separate aeration cell bottles, add 25 mL cf
laboratory pure water. Then add 0.0-, 1.0-, 2.0-, '3.0-, 4.0-, and 5.0-mL aliquots cf.
the working standard solution using Class A volumetric pipets. This corresponds to C,
200, 400, 600, 800, and 1,000 ng of Hg, respectively. Proceed with the calibration,
following steps 2 through 7 of Section 5.3.1, Recorder Adjustment. Analyze the
calibration standards by measuring the lowest to the highest standard. Be sure tc
allow the recorder pen to return fully to the baseline before the next standard is
analyzed. This step is particularly critical with peak area measurements. Repeat this
procedure on each aliquot size until two consecutive peaks agree within 3% of the
average.
Between sample analyses, place the aerator section into a 600-mL beaker
containing approximately 400 mL of water. Rinse the bottle section of the aeration
cell with a stream of water to remove all traces of the reagents from the previous
sample. These steps are necessary to remove all traces of the reducing agent between
samples to prevent the loss of Hg before aeration. It will be necessary, however, -z
wash the aeration cell parts with concentrated HC1 if any of the following conditions
occur: (I) a white film appears on any inside surface of the aeration cell; (2) the
calibration curve changes suddenly; or (3) the replicate samples do not yield
reproducible results.
- o
-------
Section No. 3.19.5
Date September 3, 1992
Page 11
Recorder or integrator responses should be documented on the analytical data
form for Calibraxon Standards (Figure 5.4) . Subtract the average peak height (or peak
area) of the measurement blank (0.0-mL aliquot) -which should be less than 2% of
recorder full scale— from the averaged peak heights of the 1.0-, 2.0-, 3.0-, 4. 0-, and
5.0-mL aliquot standards. If the blank absorbance is greater than 2% of full-scale,
the probable cause is Hg contamination of a reagent or carry-over of Hg from a
previous sample. Plot the corrected peak height of each standard solution versus the
corresponding total Hg mass in the aeration cell (in ng) .
Calculating the measured standard Hg mass (P) may be performed in two ways:
a linear regression program provided by a hand calculator (or other computing device)
or the manual least squares method described below: . . ..
P = (S) (Y) Equation 5-1
where :
Y = Peak height or integrator response, mm or counts.
S = Response factor, ng/mm or ng/counts (from Equation 5-2) .
-*
and > ' ""
*iy i +X2y2-Hx3y3 +x4y 4 +xsy 5+x6y 6
S = - Equation 5-2
where :
x = Standard mass value,' ng.
Complete the analytical data form for analyzing calibration standards (Figure
5.4) for each standard. Calculate the deviation of each standard measurement average
from the expected value (standard mass value), x. If the percent deviation from the
expected value is greater than 5% for any standard measurement, the calibration must
be repeated.
5.3.3 QC Operations— The quality of the analytical results can be assessed by
analyzing a variety of standard reference solutions (SRMs) of known high accuracy, such
as those available from the National Institute Standards and Technology (NIST) and
other government agencies. Standard solutions prepared by commercial suppliers that
meet NIST traceability criteria are also useful.'- If these solutions are not available,
the analysis of laboratory-prepared standard solutions may be used from a source
(supplier) other than the source of the calibration standards. These solutions will
be known as QC solutions. For example, if the calibration standards were prepared by
dilutions of a 100 mg Hg/mL solution from supplier A (or from an in-house prepared
solution from the pure mercury salt), then a QC solution might be prepared from
dilutions of one of the following:
1. An NIST Hg solution or other SRM.
-------
Section No. 3 .19.5
Date September 3, 1952
Page 12
o
Plant Location
Date Analyst
Standard
Identifier
Std 1
Std 2
Std 3
Std 4
Std 5
Std 6
Standard
mass
(x)
(ng Hg)
0
200
400
|_
600
800
1000
Integrator Response
Peak Height or Area
(y) , (mm)
1
2
,
Avg
1
..rl
1
1 ''
Measured
Standard
mass
(P)
(ng Hg)
Deviation
(%)
1
!
I
I
Equation for Linear Calibration Curve, Average Response as a function of standard
concentration.
O
y = mx + b = (
)x + (
where:
y = Instrument curve slope mm or area count =
ng Hg
x = Standard concentration (ng Hg) =
b = I = Intercept term (mm or area count) =
Measured Standard Concentration (P)
Equation 1
Equation 2
Equation 3
Equation 4
P(ng Hg) = Average Instrument Response (y) - Intercept (I)
Calibration Curve Slope
ng Hg
Ecuatior. 3
Deviation
Deviation (%) = P (nq Hq) - x (no Hq) x 100%
x (ng Hg)
Deviation = (
) x 100% =
Ecuatior. €
O
Figure 5.4. Analytical data form for analysis of calibration standards.
-------
Section No. 3.19.5
Date September 3, 1992
Page 13
2. A commercial QC solution that has been tested against an NIST solution
(or equivalent) by manufacturer A.
3. A 1,000-mg Hg/mL solution from manufacturer B.
Record analytical data for QC samples on Figure 5.5. QC solutions may be used
for a variety of analytical accuracy assessments. These include three check samples
(Check Sample A, B, and C):
A. Checks of the accuracy of the calibration operations, ChecJc Sample
A—When analyzed immediately following the calibration, the measured QC
sample value must be within 5% of the expected value described in this
section, or the calibration must be repeated. These QC samples are
known as Initial Calibration Verification (ICVs) Check samples.
B. Checks of the drift of the calibration, Check Sample B-For any of a wide
variety of conditions that may be related to instrument warmup or
instrument component deterioration, :the repeated analysis of a given
sample or standard will vary over time. To ensure thit the analysis is
"in control," a QC solution is measured at least every five samples.
If the average measured value of the QC solution has changed by more
than 10% from the expected value, the causes must be identified and
corrected. The calibration is then repeated, and all samples analyzed
since the last successful "drift"_rQC sample analysis must be repeated..
This "drift" QC sample is known as a CCV sample. It is worth noting
that the CCV need not be a "standard" type solution; any Hg-containing
solution may be used for the CCV, provided the Hg level in the aeration
cell is between 200 and 1,000 ng. Again, the measured value of this
solution must not vary more than 10% from the expected value.
C. Measuring spike recovery check sample, Check Sample C—Spiking a digested
sample (a prepared sample) with a standard solution provides a means of
assessing Hg recovery associated with the measurement process (sample
matrix effect). The steps below must be followed to determine spike
recovery:
a. After completing all sample preparation steps in Subsections 5.2.1
and 5.2.2, spike a 10-mL aliquot of Analysis Sample No. A.I with 10
mL of spiking solution containing a similar concentration of Hg, or
with 10 mL of a spike at least 10 tiroes greater than the detection
limit, whichever is greater.
b. Spike a 10-mL aliquot of Analysis Sample No. HC1 A.2 with 10 mL of
spiking solution containing a similar concentration of Hg as the
field sample, or a spike at least 10 times greater than the
detection limit, whichever is greater.
c. After all samples are analyzed, subtract the results of the spiked
and unspiked samples. If this spike is not within 15% of the expe-
cted value, then the Hg response may be owing to matrix effects. If
so, all sample digests must be analyzed by the method of standard
additions (MSA).
f\
i
v:
-------
Date samples
Plant
received Date samples analyzed
Run number (s)
Location Analyst
Calibration factor (S) Intercept (I), if applicable
QC
Sample
Number
Analysis
Number
Instrument
Response
(mm)
Deviation of replicate measurements, (%)
Mass of QC sample
without intercept
(ng Hg)
Mass of QC sample
with intercept
Mean
Instrument
Response
(mm)
Mean
Instrument
Response (Mass
Percent Blank Dilution QC
Deviation Corrected Factor Sample
(ng Hg) (y) (F) (ng Hg)
(A, - A2)
x 100
A, -f A,
2
- ( ) - ( ) 100 =
( ) + ( )
2
v o to
$i 01 (f)
- S x v x F 0 Z ~
— o-A.yJvr (liOfT
= _ „,- X X ......... = f-1 rn O
rr 2
S o
= S (y - I) F §•'
i - \ - V. w
(ng Hg)
O
Figure 5.5. Analytical data form for analysis of QC samples.
ID
K)
O
-------
Section No. 3.19.5
Date September 3, 1992
Page 15
Operations involving the use of QC samples are described in more detail below.
Note that spikes always must be measured using the linear portion of the calibration
curve (as with actual samples) . QC samples with Hg values exceeding the linear portion
of the calibration curve must be diluted and reanalyzed according to the sample
analysis procedure (Subsection 5.4.3).
Preparing the ICV Solutions—If the source of the ICV is a commercial
1,000-mg/mL stock solution, it must be diluted according to the procedure described in
Subsection 1.5.3 for intermediate and working standard solutions.
Measuring the ICV Solution-Analyze a 2- to 5-mL aliquot (i.e., 200-500 ng Hg)
of the ICV working standard solution (some mid-point aliquot). Duplicate measurements
should agree within 3% of the average. If not, determine the cause for error (consult
the laboratory supervisor if necessary), correct the problem, and recalibrate the
analysis system. Repeat as necessary. If the QC solution source is not a 1,000-mg
Hg/mL stock solution, prepare the intermediate QC solution (QC working solution) as
follows:
1. Pour about 15 mL of the solution into a clean beaker. NOTE: To avoid
contamination, do not pipet directly from the bottle.
2. Pipet (using a glass pipet of at least 5-mL volume) an appropriate
aliquot into a suitable clean glass volumetric flask, according to Table
5.2. Pipet 2 mL of the QC working solution for measurement. Use Table
5.2 to determine the expected values for the QC sample (ICV) .
Preparing and Measuring the Initial Blank Verification (IBV) and Continuing
Blank Verification (CBV)-Vlith the conventional measurement system, these verifications
"'''' may be performed merely by adding 50 mL of water, hydroxylamine sulfate solution, and
stannous chloride as described in Subsection 5.3.2.
Preparing and Measuring Spiked Sample, Check Sample C—To determine whether
there are sample matrix effects during the measurement, one sample digest must be
analyzed in the presence of added Hg. The added (spiked) Hg recovery must be within
85-115%, or the MSA must be employed for each sample and blank digest.
-------
TABLE 5.2 PREPARATION OF QC SOLUTIONS
Certitied value of
QC source solution
(ng Hg/mL)
<1
1-5
5-20
20-100
Aliquot of QC
source solution
for dilution, (mL)
A
5
5
5
Dilution
Volume, mL
vd
100
250
1000
QC working
solution
concentration
using Eq. 1
(ng Hg/mL)
cu,
Volume of
working solution
taken for
analysis, (mL)
«
2
2
2
2
Expected
value, (n
using Eq.
M
Hg
g)
2
where:
where:
Cws
cecd
vd
A
M.
M
'Hg
£std
Vd
x_A
Concentration of QC "working" solution (WSQC), ng Hg/mL.
Concentration in ng Hg/mL of QC source solution ' (QC).
Dilution volume in mL.
Aliquot of QC source solution added to volumetric flask in mL.
Cw. x VB.
Expected ng Hg in aeration flask.
Aliquot of Cw, taken for measurement, mL.
Equation 1
Equation 2
>o o en
P> P> (D
iq rr o
(K n> rr
H-
M Cfl 0
0*3
rt Z
n o
O
o
O"
-------
below:
Section 3.19.5
Date April 3, 1992
Page 17
The procedure used to determine the existence of matrix effects is described
1. Analyze an aliquot of the sample and record the sample aliquot size used
(see Subsection 5.4.3).
2. Calculate the Hg content in ng of the sample aliquot.
3. Determine a working standard aliquot size that equals or exceeds the
sample response from Step 2.
4. Add the value determined from Step 3 to an additional sample aliquot
identical to that used in Step 1.
5. Analyze this spiked sample and record the response.
6. The spike recovery is calculated as follows:
% Recovery
= C spiked sample -
C spike
C sample x 100
Equation 5-3
where;
C spiked sample
C sample
C spike
= Measured Hg in spiked sample, mg.
= Measured Hg in -onspiked sample, mg.
= Hg added to sample, mg.
Note: To ensure the validity of the spike measurement, it is imperative that
the measurement result fall within the range of the calibration.
Method of Standard Addition Analysis—If the recovery result obtained from the
section above on the measurement of spiked samples falls outside the 85-115% range,
then the MSA must be employed for all sample digest measurements. This procedure is
described below:
1. Repeat steps 1 and 2 of spiked sample measurement above to determine the
level of Hg in the sample (designated S0),r ;
2. To a second, identical sample aliquot, add a working standard volume
that contains a Hg level that is approximately 50% of the sample Hg
level. Refer to this spiked sample as Slf and record the exact aliquot
volume of sample and working standard.used.
3. Analyze spiked sample (Sj) .
4. To another, identical sample aliquot, add a working standard aliquot
that contains an Hg level that is approximately equal to that of the
sample. Refer to this spiked sample as S2, and record the exact aliquot
volume of sample and working standard used.
5. Analyze the spiked sample (S2) .
6. To another, identical sample aliquot, add a working standard aliquot
that contains an Hg level that is approximately 1.5 times that at the
sample. Refer to this spiked sample as S3, and record the exact aliquot
volume of sample and working standard used.
7. Analyze the spiked sample (S3) .
8. The peak intensity of each solution is determined and plotted on the
vertical axis of a graph. The concentrations of the known standards, are
plotted on the horizontal axis. When the resulting line is extrapolated
back to zero absorbance, the point of interception of the abscissa is
the concentration of the unknown. The abscissa on the left of the
-------
Section 3.19
Date April 3,
Page 18
1992
o
ordinate is scaled the same as on the right side, but in the opposite
direction from the ordinate. An example is shown in Figure 5.6.
u
c
ca
&
i-i
0
c/1
A
<
Zero
Absorb
Concen'
of Sa
ance s^[
/
1
f
_-
\ \
«,
r i
i Mass (ng)
Addition 0 Addn 1 Addn 2 Addn 3
-ration No addition50* Qf 100% of 150X of
mple Expected Expected Expected
Amount Amount Amount
O
STANDARD ADDITION PLOT
Figure 5.6. Method of standard additions for field samples.
To perform a valid MSA analysis, three criteria must be met:
1. The MSA standard curve must be linear using the criteria in Subsection
5.3.2.
2. The spiking level of Hg must be at least 50% of Hg in the sample.
3. The spiking level must be at least 10 times the detection lirr.it
(approximately 20 ng).
O
-------
Section 3.19.5
Data April 3, 1992
Page 19
5.3.4 Field Sample Analysis-Repeat the procedure used to establish the calibration
curve with an appropriately sized aliquot (1 to 5 mL) of the diluted sample until" two
consecutive peak heights agree as follows:
Hg mass, ng Limits (% deviation from average)
<5 50
5-15 15
15-100 ; 5
>100 3
/ An aliquot peak maximum (except the 5-mL aliquot) must be greater than 10% of
the recorder full scale. If the peak maximum of a 1-mL aliquot is off scale on the
recorder, further dilute the original source sample to bring the Hg concentration into
the calibration range of the spectrophotometer.
Run a CBV and a CCV at least after every five samples to check the
spectrophotometer calibration drift; recalibrate as necessary (see Subsection 5.3.3).
It is recommended that at least one sample from 'each stack test be checked by the
method of standard additions to confirm that matrix effects have not interfered with
the analysis (see Subsection 5.3.3). Record alKriata for field sample analysis on.the
Method 101A field analysis data form, Figure 5.7, >or similar form. • \
Analysis Samples No. A.I and HCl A.2—After sample preparation of each sample
run, two sample fractions must be analyzed for Hg to determine the total ng of Hg.
Analysis Sample No. A.I is the filtrate of the KMnO< absorbing solution and rinse and
the digestate of the glass fiber filter, if applicable. Analysis Sample No. A.I will
be 1,000 mL or more, measured to within 1 mL. Analysis Sample ,,Nor. HCl A.2 is f the
digestate of residue and; Whatman 40 filter paper and HCl rinse, if applicable.
Analysis Sample No. HCl A.2 will be 500 mL. A recommended sequence of analysis is
presented in Table 5.3.;
Analysis Samples No. A.I Blank and HCl A.2 Blank-Each test series requires
that a sample blank be taken. The sample blank is prepared in the same manner as the
field samples. The analysis of the sample blank will have the same two 'fractions as
each field sample. The blank will be analyzed at the same time and in the same manner
as the field samples, with the exception that a 10-mL aliquot shall be used for
analysis. A recommended sequence of analysis is presented in Table 5.3.
Container No. 2 (Silica Gel)4-Weigh the spent silica gel (or silica gel plus
impinger) to the nearest 0.5 g using a balance. (This step may be conducted in the
field.)
-------
Date samples received
Date samples analyzed
Plant Run number (s)
Location Analyst
Calibration factor (S) Intercept (I), if applicable
Field
Sample (Analysis
Number
,
Number
Instrument
Response
(mm)
Mean
Instrument
Response
(mm)
(A,
Deviation of replicate measurements. (%) =
A,
_ _£_
(
Percent
Deviation
(ng Hg)
.
Mean
Instrument
Response
Blank
Corrected
(y)
JMass
Dilution
Factor
(F)
Field
Sample
(ng Hg)
- A2)
x .100
+ A,
2
) - ( ) 100 =
) 4 ( )
2
Mass of QC sample
without intercept
(ng Hg)
Mass of QC sample
with intercept
(ng Hg)
= S x y x F
x
O
= S (y - I) F
Figure 5.7. Analytical data form far analysis of field samples.
•ti a tn
pi fu ID
id rr n
eo (o rt
H-
to > o
0-0 3
10
VO
(O
O
-------
Section 3.19.5
Date April 3, 1992
Page 21
TABLE 5.3 RECOMMENDED ANALYTICAL SEQUENCE"
Sequence No.
1
2
3
4
5
6
7
8
9
10
J) "
12
13
14
15
16
17
18
19
20
Sample ID
IBV
repeat
ICV
repeat
ccvb
repeat
A.I blank
repeat
HC1, A. 2 blank
repeat
A.I, Run 1
repeat
A.I Spike, Run le
repeat
HC1 A. 2 Run ,lc „,
repeat*3
CCV
repeat
CBV
repeat
Sequence No.
21
22
23
24
25
26
27
28 \
\
29
30
31
32
33
34 ^
' ::" 35 ~; . -,
Sample ID
HC1 A. 2, Run 1 spike
repeat
A.I, Run 2
repeat
HC1 A. 2, Run 2 '
repeat
A.I, Run 3
repeat
HC1 A. 2, Run 3
repeat
CCV
repeat
CBV
repeat
Repeat Calibration"1 '
"Assuming a valid calibration has been performed.
blf different-than ICV. ,t, ,
^Any A.I spike from runs 1, 2, or 3. ,-.-
this point, if the recovery is 85 to 115%, proceed to Step 26; if
,ot, all samples must be run /using MSA (Subsection 5.3.3).
-------
Section 3 .19 . 5
Date April 3, 1992
Page 22
5.4 Alternate Analytical Apparatus
Alternative systems are allowable as long as they meet the following criteria:
5.4.1 Measurement Technique—The system is based on cold vapor atomic absorption
techniques.
5.4.2 Analyte Recovery—Eighty-five-115% of the spike is recovered when an aliquot
of a source sample is spiked with a known concentration of Hg (II) compound.
5.4.3 Calibration Curve—A linear calibration curve is generated and two consecutive
standards of the same aliquot size and concentration agree within the following limits.
Hg mass, ng/mL Limits (% deviation from average)
10 3
5.4.4 Sensitivity—The system is capable of detecting 0.2 ng Hg/mL for flow-injection
systems or 20 ng Hg for batch systems.
o
•O
i--u«^>—<
An example of a flow-injection analytical system is depicted in Figure 1
Note that these systems inject samples in a semicontinuous manner; consequently, the
solution concentration is monitored, not the total Hg in the entire sample. Therefore,
the total Hg in a given sample digest is calculated as follows:
MHa = CHg x V Equation 5-4
where:
MHg = Total mg Hg in each sample digest from Section 5.2.
CHg = Measured concentration in mg Hg/mL.
V = Total volume in milliliters of the sample digest.
These calculations are shown in Section 3.19.6.
5.4.5 Data Quality Assessment for Alternate Analytical Systems-QC solutions used to
determine the accuracy of the calibration may be measured directly without performing
the calculations described in Subsection 5.3.3. This procedure, of course, is based
on the assumption that the sample concentration value does not exceed that of the
highest calibration standard, thereby requiring a dilution.
O
-------
follows:
Section 3.19.5
Date April 3, 1992
Page 23
Determining matrix effects on the measurement recovery is performed as
1. Determine the Hg concentration in the sample digest.
2. Remove two 10-mL aliquots of the digest and place in clean 20-mL
beakers.
3. To one aliquot, add 1 mL of distilled deionized water and mix by
swirling the beaker (S0) . — -
4. To the other aliquot, add 1 mL of a standard that is 10 to 20 times the
solution concentration of the sample; mix the beaker contents (S^ .
5. Measure both solutions for Hg content.
6. The recovery of the added spike is as follows:
% Recovery
Ms, - Mso
X 100
Equation 5-5
where:
MSO
= mg Hg in spiked sampler
\
= mg Hg/mL in St x 11 mL.
= mg Hg in sample spiked with water.
= mg Hg/mL (of spiking solution) x 1 mL.
The recovery should'be between 85-115%. Otherwise, the method of additions
must be employed for each sample of the sample run. --
Method of Additions—In this method, equal volumes of sample are added to a DI
water blank and to three standards containing different known amounts of the test
element. The volume of the blank and the standards must be the same. The absorbance
(peak height, counts, etc.) of each solution is determined and then plotted on the
vertical axis of a graph. The concentration of the known standards are plotted on the
horizontal axis. When the resulting line is extracted back to zero absorbance, the
point of interception of the abscissa is the concentration of the unknown. The
abscissa on the left of the ordinate is scaled the same as on the right side, but in
the opposite direction from the ordinate. An example is shown in figure 5.6. .......
5.5
Posttest Checklist
Posttest checklists for QC sample analysis and field sample analysis are
presented in figures 5.8 and 5.9. .
-------
Section 3.19.5
Date April 3, 1992
Page 24
o
QC Sample Analysis Checklist
Date Plant Name Sampling Location
Calibration Standards and Matrix Checks
Mercury Stock Solution, 1 rag Hg/mL:
Prepared in-house'? (Y/N)
Source of mercury (II) chloride
Commercial stock solution? (Y/N)
Source
Intermediate Mercury Standard Solution, 10 mg/mL:
Date prepared
Used glass pipet? (Y/N)
Source and grade of HNO3
Working Mercury Standard Solution, 200 ng Hg/mL:
Prepared today? (Y/N)
Used glass pipet? (Y/N)
Calibration Standards:
mL of working standards volume of volumetric flask, mL
#1
#2
#3
#4
#5 •
#6
#7
Instrumentation:
Spectrophotometer type
O
Moisture Removal System:
Optical cell heating system? Calibrated?
Moisture trap used? What type?
Data Recording System:
Recorder Integrator Other
Describe
Peak height Peak area
O
Figure 5.8. QC sample analysis.
-------
Section 3.19.5
Date April 3, 1992
Page 25
Cold Vapor Generation System:
Standard batch system?
Alternate system?
Describe alternate system?
Aeration gas .... Aeration gas flow
Gas cylinder? . Peristaltic pump?
Standardization:
Glass pipets used? _
mL of . Standard value* ... ?.••*:;•?. .»• •; .
working standard .... -*(hg) ' •"••"•' Reading 1 ;': Reading 2 %Difference
"If using an alternate system that uses flow injection this value may be expressed as
concentration, e.g., |ig/L, ng/L, etc.
Calibration coefficient - -
Offset at origin (measured response of calibration blank) ng or % of scale.
Initial Calibration Verification (ICV) ; "' '
QC check sample source
Certified or expected concentration
Measured concentration
% Difference ''"' • .;'
Initial Calibration Blank Verification (IBV);
Measured value
Below detection limit?
Matrix Interference Check:
Method of additions performed for one test site sample?
Spike added
Spike recovered
% recovery = Spiked sample value - unspiked sample value =
spike value
Figure 5.8. (Continued)
-------
Section 3.19.5
Date April 3, 1992
Page 26
o
If the recovery was outside of 85-115%, were samples run using the method of standard
additions?
Describe ..
Continuing Calibration Verification (CCV) - Check sample of standard to be reanalyzed
after every five samples:
Standard used (source) .
Expected value/unit
Was measured value/unit always within 10% of expected value?
Final Standardization:
Glass pipets used?
mL of Standard value"
working standard (ng) Reading 1 Reading 2 %Difference
O
"Alternate analytical systems may express Hg value as a concentration (e.g., mg/L Hg).
Calibration coefficient
Offset at origin ng or % of scale.
Figure 5.8. (Continued)
O
-------
Section 3.19.5
Date April 3, 1992
Page 27
Sample Analysis Checklist " ~" '
Date • Time ._ Operator ••• Observer _______^___
Sample Analysis
Were all sample digests analyzed within 48 h of preparation? _____( Y/N)
Were 10 mL of samples A.I blank and HC1 A. 2 blank used as a minimum?
_ (Y/N)
Were duplicate measurements performed as a minimum for all blank and sample digests?
Did duplicate measurements meet the "percent difference" criteria outlined in Table 5.2
_ (Y/N)
Was the largest possible aliquot (20 mL) used when a measurement was below the
detection limit? _ (Y/N)
If a sample measurement exceeded the highest calibration standard, were appropriately
smaller aliquots always taken to ensure that results fell within the calibration range?
_ (Y/N)
If 1-mL aliquots taken for measurement still were off scale, were sample digests
diluted so that results were within the linear range of the standards? (Y/N)
What volumetric glassware (pipets) was used to add sample digests to the aeration
flasks? _ (mL) '
What volumetric glassware (pipets/ flasks) was used to dilute sample digests? _______
__ (if necessary) ,
If the calibration check samples differed by greater than 10% of the expected values,
was the system recalibrated? (Y/N)
Were CBV and CCV samples analyzed after ever five samples? _ (Y/N)
Were all samples run after the previously CBV and CCV sample analyses? . --
_ (Y/N) '
Was the full standardization performed at the end of the sample analysis? _ (Y/N)
Figure 5.9. Method 101A sample analysis checklist.
<
-------
Section 3.19 .5
Date April 3, 1992
Page 28
o
TABLE 5.1 ACTIVITY MATRIX FOR SAMPLE ANALYSIS
Characteristic
Sample
preparation
All
calibrations
1
1
(Acceptance limits
1
| Samples and
(blanks prepared
I under same
| conditions
1
1
|(1) Reagents and
(volumes used
(during
(measurement of
| samples and
Frequency & method
of measurement
Dilute samples so
that matrix
concentrations are
are identical to
original sample
(Action if
| requirements
| are not met
I
(Adjust
(dilutions, if
(possible;
(otherwise report
(to Administrator
1
| Reanalyze
| samples
j
1
1
(standards are (digest |
| identical
1
| (2) Perform
.— • ' '
•-
Prepare fresh
| 6 -point (daily
(calibration curve
I including
(calibration blank
1
I (3) Calibration
(coefficient
(better than 0.999
'
Each calibration
point is the
average of
( (duplicate
1
1
1
| Prepare fresh
(daily
1 /
1 (
1
(Repeat
(calibration
1
1
| (measurements |
1 1
Calibration
Verification
Check Samples
(ICV)
Calibration
Blanks
(Analysis within (Analyze after
| 5% of expected or
I certified value
1
1
every calibration
(Must be below (Analyze after
(detection limit
Verification I
(IBV)
Continuing
Calibration
Verification
Sample
(CCV)
1
I
I
every
calibration
(Must be within (Analyze after
J10% of expected
(value
1
|
1
every 5th sample
(Ensure quality
(of check sample
| or repeat
| calibration
(check for
(potential
(contamination
(and repeat
| calibration
1
| Repeat
| calibration
(and repeat all
(samples since
| last successful .,
|CCV analysis v __ {
O
o
(Continued)
-------
Section 3.19.5
Date April 3, 1992
Page 29
TABLE 5.1 (Continued)
Characteristic
Continuing
Blank
Verification
(CSV)
Matrix check
sample
Duplicate
measurements
Data recording
Acceptance limits
Must be below
detection limit
Recovery of
sample digest
spike 85-115%
See Subsec.
5.3.2
All pertinent
data recorded
on figs. 5.1, 5.2
Frequency & method
of measurement
Analyze after
every 5th sample
One sample digest
from every stack
test is spiked at
a level at least
equal to sample
digest
conccntrat ion
All standard and
sample analyses
Visually check
(Action if
| requirements
(are not met
1
1 Repeat
(calibration
land repeat all
(samples since
I last successful
|CBV analysis
!
(Analyze all
(samples using
|the method of
| standard
(additions
1
1
(Repeat until
| agreement is
| achieved
1
(Supply missing
(data
1
-------
o
o
o
-------
Section No. 3.19.6
Date September 3, 1992
Page 1
6.0
CALCULATIONS
Calculation errors resulting from procedural or mathematical mistakes can be
a part of total system error. Therefore, it is recommended that each set of calcula-
tions be repeated or spotchecked, preferably by a team member other than the one who
performed the original calculations. If a difference greater than typical round-off
error is detected, the calculations should be checked step-by-step until the source of
error is found and corrected.
Calculations should be carried out to at least one extra decimal figure beyond
that of the acquired data and should be rounded off after final calculation to two
significant digits for each run or sample. All rounding of numbers should be performed
in accordance with the ASTM 380-76 procedures.
A computer program is advantageous in reducing calculation errors. If a
program is used, the original data entered should be included in the printout for
review. If differences are observed, a new computer run should be made. A computer
program also is useful in maintaining a standardized format for reporting results. The
data shown will allow auditing the calculations.
Table 6.1 at the end of this section summarizes the QA activities for
calculations. -*~ .
In the next section, nomenclature and equations have been divided into two
groups. The first group (Section 3.19.6.1 to 3.19.6.4) deals with sampling calcula-
tions. The second group (Section 3.19.6.5 to 3.19.6.13) deals with analytical and
emissions calculations.
6.1
Sampling Nomenclature from Method 5
An
Bw,
I
L.
L,
R
Cross-sectional area of nozzle, m2 (ft2).
Water vapor in the gas stream, proportion by volume.
Percent of isokinetic sampling.
Maximum acceptable leakage rate for either a pretest leak check or
for a leak check following a component change, equal to 0.00057
mVmin (0.02 cfm) or 4% of the average sampling rate, whichever is
less.
Individual leakage rate observed during the leak check conducted
prior to the "ith" component change (i = 1, 2, 3...n), mVrnin (cfm).
Leakage rate observed during the posttest leak check, mVmin (cfm).
Molecular weight of water, 18.0 g/g-mole (18.0 Ib/lbmole) .
Barometric pressure at the sampling site, mm Hg (in. Hg).
Absolute stack gas pressure, mm Hg (in. Hg).
t,
Standard absolute pressure, 760 mm Hg (29.92 in. Hg) .
Ideal gas constant, 0.06236 [ (mm Hg) (m3) ] / [ (°K) (g-mole) ] {21.85 [(-
in. Hg) (ftJ)]/H°R) (lb-mole)]}.
-------
Section No. 3 .19.6
Date September 3, 1992
Page 2
Tm = Absolute average DGM temperature, °K (°R).
T. = Absolute average stack gas temperature, °K (°R).
T,ld = Standard absolute temperature, 293 °K (528R).
V,c = Total volume liquid collected in impingers and silica gel, mL.
Vm = Volume of gas sample as measured by dry-gas meter, dcm (dcf).
vm.tdi = Volume of gas sample measured by the dry-gas meter, corrected to
standard conditions, dscm (dscf).
Vwi.cdi = Volume of water vapor in the gas sample, corrected to standard
conditions, scm (scf).
vs = Stack-gas velocity, calculated by Method 2, Equation 2-9, using
data obtained from Method 5, m/s (ft/s).
Y = Dry-gas meter calibration factor.
AH = Average pressure differential across the orifice meter, mm H20 (in.
H20).
pw = Density of water, 0.9982 g/mL (0.002201 Ib/mL).
6 = Total sampling time, min.
o
o
6j = Sampling time interval, from the beginning of a run until the first
component change, min.
6, = Sampling time interval, between two successive component changes,
beginning with the interval between the first and second changes,
min.
6P = Sampling time interval, from the final (nth) component change until
the end of the sampling run, min.
13.6 = Specific gravity of mercury.
60 = S/min.
100 = Conversion to %.
O
-------
6.2
6.3
6.4
Section No. 3.19.6
Date September 3, 19S2
Page 3
Conversion Factors
From
scf
g/ft3
g/ft3
g/ftj
To
m3
gr/ft3
lb/ft3
g/m3
Multiply by
0.02832
15.43
2.205 x 10°
35.31
Average Dry-Gas Meter Temperature and Average Orifice Pressure Drop
See data sheet (Figure 4.1).
Dry-Gas Volume
Correct the sample volume measured by the dry-gas meter to standard conditions
(20 °C, 760 mm Hg or 68 °F, 29.92 in. Hg) by using Equation 6-1.
^ AH ,
V=VmY
V = K, Vu Y
Equation 6-1
where:
Kj = 0.3858 °K/mm Hg for metric units.
• = 17.64 °R/in Hg for English, units.
Note: Equation 6-1 can be used as written unless the leakage rate observed
during any of the mandatory leak checks (i.e., the posttest leak check or leak-checks
conducted prior to component changes) exceeds La. If Lp or L, exceeds La, Equation 6-1
must be modified as follows:
(a) Case I. No component changes made during sampling run. In this case, replace Vj
in Equation 6-1 with the expression:
[V, - (L. - La) 6]
-------
Section No. 3 .19 .6
Date September 3, 199
Page 4
o
(b) Case II. One or more component changes made during the sampling run. In this
case, replace Vm in Equation 6-1 by the expression:
(V. - (L, - La) 81 - (L, -La) 6, - (Lp - L.) 8p]
1-2
and substitute only for those leakage rates (L, or Lp) that exceed La.
6 . 5 Volume of Water Vapor
Vw(.td, = vic ^V^ = K2 Vle Equation 6-2
l-Lu CBtd
where :
K2 = 0.001333 rn'/mL for metric units.
= 0.04707 ftVmL for English units":"
6.6 Moisture Content
B = V"'stdl Equation 6-3
us 17 nTlT
vm(std)
O
Note: In saturated or water droplet-laden gas streams, calculate the moisture
content of the stack gas in two ways: from the impinger analysis (Equation 6-3) and
from the assumption of saturated conditions. The lower for B,,. shall be considered
correct. The procedure for determining the moisture content based upon assumption of
saturated conditions is given in the Note of Section 1.2, Method 4. For the purposes
of this method, the average stack-gas temperature from Figure 4.2 may be used to make
this determination, provided that the accuracy of the in-stack temperature sensor is
± 1 °C (2 °F).
6.7 Nomenclature from Method 2
A = Cross-sectional area of stack, m2 (ft2).
BU5 = Water vapor in the gas stream (from Method 5 or Reference Method
4), proportion by volume.
Ct. = Pitot tube coefficient, dimensionless.
K,, = Pitot tube constant - 34.97 for the metric system and 85.49 for
the English system.
Mj = Molecular weight of stack gas, dry basis (see Section 3.1
g/g-mole (Ib/lb-mole).
Mf = Molecular weight of stack gas, wet basis, g/g-mole (Ib/lb-mole).
-------
Section No. 3.19.6
Date September 3, 1992
Page 5
= Ma (1 - BW9) + 18.0 Bw,
Pb.r = Barometric pressure at measurement site, mm Hg (in. Hg) .
Pg = stack static pressure, mm Hg (in. Hg) .
P. = Absolute stack pressure, mm Hg (in. Hg) ,
D
6.8
P.td
Q.d
TBld
vs
Ap
3,600
18.0
= Standard absolute pressure, 760 mm Hg (29.92 in. Hg) .
= Dry volumetric stack gas flow rate corrected to standard condi
tions, dsmVh (dscf/h) .
= Stack temperature, °C (°F) .
= Absolute stack temperature, °K (°R).
= 273 4- t, for metric. -*"
= 460 -*• t. for English.
= Standard absolute temperature, 293 °K (528 °R) .
= Average stack gas velocity, m/sec (ft/s) .
= Velocity head of stack gas, mm H20 (in. H20) .
= Conversion factor, s/h.
= Molecular weight of water, g/g-mole (Ib/lb-mole) .
Average Stack Gas Velocity
V *
CP <
Ts (avg)
Equation 6-4
-------
6.9
6.10
Average Stack Gas, Dry Volumetric Flow Rate
Q = 3600 (l-Bus) vsA
Isokinetic Variation
6.10.1 Calculation from Raw Data-
100 T
I =
V2c + (Vm -
/n
60 fc) vs
Section No. 3.19.6
Date September 3, 1952
Page 6
o
Equation 6-5
Equation 6-6
where:
6.10.2
K3 = 0.003454 [(mm Hg) (m3)]/[(mL) (°K)] for metric units.
-*"*
= 0.002669 [(in. Hg) (ft3)]/[(mL)(°R)], for English units.
Calculation from Intermediate Values-
J =
T , v
•'•std vs
\n Pe 60 (1 - Bus)
Equation 6
-O
= K,
v
's ^m(std)
Equation 6-E
where:
6.11
K4 = 4.320 for metric units.
= 0.09450 for English units.
Acceptable Results
If 90%
-------
Section No. 3.19 .6
Date September 3, 1992
Page 7
6.12
Method...1.Q 1A Calculations
6.12.1 Determining Compliance-Each performance test consists of three repetitions of
the applicable test method. For the purpose of determining compliance with an
applicable national emission standard, use the average of the results of all
repetitions.
6.12.2 Total Hg-For each source sample, correct the average maximum absorbance of the
two consecutive samples whose peak heights agreed within 3% of their average for the
contribution of the blank. Then calculate the total Hg content in \ig in each sample.
Correct for any dilutions made to bring the sample into the working range of the
spectrophotometer. -•'-
m
[C
(HCl)Hg
(HCl}Hg
DF]
where:
m
(HCUH9
•(HCl blklHg
DF
DFh
[C
(HCl blklHg
DFblk]
S
10
-3
Equation 6-9
blk
Total blank corrected Jig of Hg in HCl rinse and HCl digestate
of filter sample.
Total ng of Hg analyzed in the aliquot from the 500-mL Analysis
Sample No. HCl A.2. :
Total ng of Hg analyzed in aliquot of.the 500-mL Analysis Sample
No HCl A.2 blank.
Dilution factor for the HCl digested Hg-containing solution,
Analysis Sample No. HCl A.2. This dilution factor (DF) applies
only to the intermediate dilution steps because the original
sample volume (Vf)HC1 of HCl A.2 has been factored out in the
equation, along with the sample aliquot, (S) . In Equation 6.9,
the sample aliquot, S, is introduced directly into the aeration
cell for analysis according to the procedure outlined in Section
3.19.5.3.4. A dilution factor is required only if it is
necessary to bring the sample into the analytical instrument's
calibration range. If no dilution is necessary, then DF equals
1.0.
Dilution factor for the HCl digested solution, Analysis Sample
No. HCl A.2 blank. (Note: Normal dilution factor calculations
apply here.)
Solution volume of original Analysis Sample No. HCl A.2 andkHCl
A.2 blank, 500 ml, for samples diluted as described in Section
5.2.2.4 of this document.
10'
= Conversion factor, ug/ng.
-------
Section No. 3.19. 6
Date September 3, 1BS2
Page 8
S = Aliquot volume of sample added to aeration cell, mL.
Sb)k = Aliquot volume of blank added to aeration cell, mL.
Note: The maximum allowable blank subtraction for the HC1 is the lesser of the
two following values: (1) the actual blank measured value (Analysis Sample No. HC1 A.2
blank); or (2) 5% of the Hg content in the combined HC1 rinse and digested sample
(Analysis Sample No. HC1 A.2).
o
'imriHj
V
[([ltr].
where:
blk)Hg
'
10
-3
yblk
Equation 6-1C
(f UrlHg
-If llrlHg
Mfltr blklHg
= Total blank corrected (Jg "of Hg in KMn04 filtrate and HN03 di-
gestion of filter sample.
= Total ng of Hg in aliquot of KMnO4 filtrate and HN03 digesticr.
of filter analyzed (aliquot of Analysis Sample A.I).
-o
= Total ng of Hg in aliquot of KMnO< blank and HNO, digestion cf
blank filter analyzed (aliquot of Analysis Sample No. A.I
blank).
Vt((Url = Solution volume of original sample, normally 1000 mL fcr
samples diluted as described in Section 7.3.2 of Method 101A.
V,lblkl = Solution volume of blank sample, 1000 mL for samples diluted as
described in Section 7.3.2 of Method 101A.
Note: The maximum allowable blank subtraction for the HC1 is the lesser of the two
following values: (1) the actual blank measured value (Analysis Sample No. A.I blank;;
or (2) 5% of the Hg content in the filtrate (Analysis Sample No. A.I).
O
-------
Section No. 3.19.6
Date September 3, 1992
Page 9
Equation 6-11
where:
mHg = Total blank corrected Hg content in each sample, fig.
m(HciiH3 ~ Total blank corrected \ig of Hg in HC1 rinse and HC1 digestate of
filter sample.
mmtriHs = Total blank corrected Hg of Hg in KMnO4 filtrate and HNO3 di-
gestion of filter sample.
6.12.3 Mercury Emission Rate-Calculate the Hg emission rate R in g/day for continuous
operations using Equation 101A-6 in Method 101A. For cyclic operations, use only the
time per day each stack is in operation. The total Hg emission rate from a source will
be the summation of results from all stacks.
where:
A.
86,400
ID'6
wist.
T.
PS
K
K
R = K
(86,4^0 x 10'6)
Equation 6-12
Total blank corrected Hg content in each sample, ug.
Average stack gas velocity, m/s (fps).
Stack cross-sectional area, m2 (ft2).
Conversion factor, s/day.
Conversion factor, g/Jig.
Dry-gas sample volume at standard conditions, corrected for
leakage (if any), m3 (ft3).
Volume of water vapor at standard conditions, m3 (ft3) .
Absolute stack-gas temperature, °K (°R).
Absolute stack-gas pressure, mm Hg (in. Hg).
0.3858 °K/mm Hg for metric units.
17.64 °R/in. Hg for English units.
-------
Section No. 3.19 .6
Date September 3, 1952
Page 10
6.13
Determining Compliance
o
Each performance test consists of three repetitions of the applicable test
method. For the purpose of determining compliance with an applicable national emission
standard, use the average of the results of all repetitions.
6.14
Hq Calculation for Alternate Analytical Systems
For alternate analytical systems, in which Hg is measured as a concentration
(mg Hg/L of sample) the Hg in mg (m^) in the original solution is calculated as
follows:
where:
CH
DF
V,
= CHg x (DF) x (Vf)
Equation 6-12
= Measured concentration of Hg in mg Hg/L of digested sample.
= Dilution factor for the Hg-containing solution used to ensure measured
sample values were, within the-c~def ined portion of the calibration
curve. i
= Solution volume of sample prepared in L.
O
TABLE 6.1 ACTIVITY MATRIX FOR CALCULATION CHECKS
Apparatus
|Acceptance
|limits
I
|Frequency and method
|of measurement
|Action if
requirements
are not met
Analysis
data form
|All data and
|calculations are
I shown
(Visually check
I
I
Complete the
missing data
Calculations
(Difference
|between check and
(original
|calculations
|should not exceed
(round-off error
(Repeat all
(calculations
(starting with raw
|data for hand
|calculations; check
|all raw data input
|for computer
(calculations; hand
(calculate one sample
(per test
Indicate errors
on calculation
form
O
-------
Section No. 3.19.7
Date September 3, 1992
Page 1
7.0 MAINTENANCE
The normal use of emission testing equipment subjects it to corrosive gases,
extremes in temperature, vibration, and shock. Keeping the equipment in good operating
order requires knowledge of the equipment and a program of routine maintenance
performed quarterly or after 1000 ft3 of operation, whichever is greater. In addition
to the quarterly maintenance, cleaning pumps and metering systems annually is
recommended. Maintenance procedures for the various components are summarized in Table
7.1 at the end of this section. The following procedures are not required, but they
are recommended to increase the reliability of the equipment.
7.1 Sampling Equipment
7.1.1 Pump-Several types of pumps may be used to perform Method 101A; the two most
common are the fiber vane pump with in-line oiler and the diaphragm pump. The fiber
vane pump requires a periodic check of the oiler jar. Its contents should be translu-
cent; the oil should be changed if not translucent. Use the oil specified by the
manufacturer. If none is specified, use SAE-10 nondetergent oil. Whenever a fiber
vane pump starts to run erratically, or during the yearly disassembly, the head should
be removed and the fiber vanes changed.
The diaphragm pump requires little maintenance. If the diaphragm pump leaks
or runs erratically, it is normally due to a bad diaphragm or malfunctions in the
{waives; these parts are easily replaced and should be cleaned annually by complete
disassembly of the train.
7.1.2 Dry-Gas Meters—The dry-gas meter should be checked for excess oil and
component corrosion by removing the top plate every 3 months. The meter should be
disassembled, and all components should be cleaned and checked more often if the dials
show erratic rotation or if the meter will not calibrate properly.
7.1.3 Inclined Manometer-The fluid should be changed when it is discolored or
contains visible matter and when it is disassembled yearly. No other routine
maintenance is required because the inclined manometer is evaluated during the leak
checks of both the pitot tube and the entire meter box.
7.1.4 Sampling Train-All remaining sample train components should be visually
checked every 3 months, and they should be completely disassembled and cleaned or
replaced yearly. Many of the items, such as quick disconnects, should be replaced only
when damaged.
7.2 Analytical Instruments
7.2.1 Spectrophoto/neter-Consult the manufacturer's operation manual for specific
maintenance activities.
7.2.2 Peristaltic Pump TuJbing-Inspect pump tubing daily. The tubing should not have
flat spots where it has contacted the pump rollers and should feel flexible. Replace
ubing if this is not the case. ^ ^
7.2.3 Desiccant-If a moisture trap is used instead of a heated optical cell, the
desiccant should be replaced daily. Both tube ends should be filled with glass wool;
the dessicant roust not be packed too tightly.
-------
Section No. 3.19.7
Date September 3, 1992
Page 2
7.2.4 Optical Cell—The windows of the optical cell should be inspected daily for an;
dust, dirt, or grease that will degrade light throughput and overall analytical
performance. Wash gently with detergent and rinse well. Dry by blotting with a towel
and wipe, if necessary, with lens paper only.
7.2.5 Spectre-photometer Windows—The windows of the spectrophotometer must also be
inspected (at least weekly) and cleaned as described in section above.
7.2.6 Tygon Connecting Tubing-Connection tubing must be inspected on a daily basis
(or more frequently) for condensation or dirt. Replace if
necessary. The existence of moisture after the dessicant trap (if used)
indicates that the dessicant needs replacing. Refer to Section 7.2.3.
o
o
o
-------
Section No. 3.19.7
Date September 3, 1992
Page 3
TABLE 7.1 ACTIVITY MATRIX FOR EQUIPMENT MAINTENANCE CHECKS
-------
Section No. 3.19.7
Date September 3, 1992
Page 4
o
TABLE 7.1 Checks (Continued)
Apparatus
Analytical
Instruments
Spectro-
photometer
Peristaltic
pump tubing
Desiccant
Optical
cell
Spectro-
photometer
windows
Tygon
connecting
tubing
Acceptance
limits
1
See owner' s
manual
'
Flexible; no
flat spots
Fresh or dry
used silica gel;
no moisture
Clean of dust,
dirt, grease,
etc.
Same as above
No condensation
or dirt
Frequency and method
of measurement
. J .•;:••'
See owner's manual ,
manual
Visually inspect
tubing daily
Inspect daily
Inspect,-daily
:
Inspect weekly
Inspect daily
Action if
requirements
are not met
See owners
Replace
Replace
Clean gently
with detergent;
rinse; blot with
towel
Same as above
Replace
O
-------
Section No. 3.19.8
Date September 3, 1992
Page 1
8.0 AUDITING PROCEDURES
An audit is an independent assessment of data quality. Independence is
achieved when the persons performing the audit apply standards and equipment different
from the standards and equipment of the regular field team. Routine QA checks by a
field team are necessary to generate quality data, but they are not part of the
auditing procedure. Table 8.1 at the end of this section summarizes the QA functions
for auditing.
One performance audit is recommended when testing for compliance with National
Emission Standards for Hazardous Air Pollutants (NESHAPs), with New Source Performance
Standards (NSPS), and as required by other government agencies. A performance audit
is recommended when testing for other purposes; and two other performance audits are
recommended. The three performance audits are:
1. An audit of the analysis of Method 101A is recommended for NESHAPs. The
use of an NIST-traceable control sample is recommended for NSPS testing
and for other purposes.
2. An audit of the sampling is suggested by Method 101A and is recommended
by the QA Handbook.
3. An audit of the data processing i,« also recommended.
It is suggested that & systems audit be conducted as specified by the QA
coordinator in addition to these performance audits. The two performance audits and
the systems audit are described in detail in Subsections 8.1 and 8.2, respectively.
8.1 Performance Audits
Performance audits are conducted to evaluate quantitatively the quality of
data produced by the sampling, analysis, or the total measurement system (sample
collection, sample recovery, sample analysis, and data processing).
8.1.1 Performance Audit of Method 101A Analysis—A performance audit for Method 101A
analysis is recommended for NESHAPs and NSPS testing using a control sample that is
NIST-traceable. Although the control sample values are known to the analyst, the
successful analysis of a control sample, as described in Subsection 5.3.3, makes the
results traceable to an NIST standard.
8.1.2 Performance Audit of the Field Test—The dry-gas meter calibration should be
checked by one of the two techniques shown below (meter orifice check or critical
orifice check).
Meter Orifice Check—Using the data obtained during the calibration procedure
described in Section 5.3, determine the AH, for the metering system orifice. The AH,
is the orifice pressure differential in units of in. H20 that correlates to 0.75 cfm
of air at 528 °R and 29.92 in. Hg. The AHg is calculated as follows:
= " §__ Equation 8-1
AH = Average pressure differential across the orifice meter, in. H20
-------
Section No. 3.19.8
Date September 3, 1992
Page 2
Tm = Absolute average DGM temperature, °R.
Pt»r = Barometric pressure, in. Hg.
6 = Total sampling time, min.
Y = DGM calibration factor, dimensionless.
VK = Volume of gas sample as measured by DGM, dcf.
Before beginning the field test (a set of three runs usually constitutes a
field test), operate the metering system (i.e., pump, volume meter, and orifice) at the
AH, pressure differential for 10 min. Record the volume collected, the DGM tempera-
ture, and the barometric pressure. Calculate a DGM calibration check value, Yc, as
follows:
o
TTJT19 Tm
• bar
Equation 8-2
where:
Yc = DGM calibration check value, dimensionless.
10 = Run time, min.
0.0319 = (0.0567 in Hg/°R)(0.75 cfm)2.
Compare the Yc value with the dry-gas meter calibration factor, Y, tc
determine that: 0.97Y
-------
Vcrlatd) = K'
Y = V
Section No. 3.19 .8
Date September 3, 1992
Page 3
Equation 8-4
Equation 8-5
where:
vcristdi = Volume of gas sample passed through the critical orifice,-—correct-v
ed to standard conditions,*dscm (dscf).
K' = 0.3858 °K/mm Hg for metric units ... • •
= 17.64 °R/in Hg for English units.
7. Average the DGM calibration values for each of the flow rates. The
calibration factor, Y, at each of the flow rates should not differ by
more than ± 2% from the average. . ...
8.1.3 Performance Audit of Data Processings-Calculation errors are prevalent when
processing data. Data processing errors can be determined by auditing the recorded
data on the field and laboratory forms. The original and audit (check) calculations
should agree within round-off error; if not, all, of the remaining data should be
checked. Data processing also may be audited by requiring that the testing laboratory
rovide an example calculation for one sample,run. This example calculation will
nclude all the calculations used to determine the emissions based on the raw field and
laboratory data. . :
8.2
System Audit
A system audit is an on-site, qualitative inspection and review of the total
measurement system. Initially, a system audit is recommended for each enforcement
source test, defined here as a series of three runs at a source.
The auditor should have extensive background experience with source sampling
or source test observation, specifically with Method 101A or Method 5. The. auditor-'•&•
functions are summarized below:
1. Observe procedures and techniques of the field team during sample
collection and sample recovery.
2. Check/verify records of apparatus calibration checks and QC used in the
laboratory analysis. . . ...
While on-site, the auditor observes the source test team's overall
performance, including the following specific operations:
1. Setting up the sampling system and checking the sample train and pitot
tube for leaks.
2. Collecting the isokinetic sampling.
3. Conducting the final leak checks.
4. Sample documentation procedures, sample recovery, and preparatipn of the
samples for shipment.
Figure 4.3 in Section 3.19.4 is a suggested field observation checklist for 101A
sampling and sample recovery, and Figure 5.9 in Section 3.19.5 is a suggested checklist
or 101A sample analysis. "• -
-------
Section No. 3.19.8
Date September 3, i992
Page 4
TABLE 8.1 ACTIVITY MATRIX FOR AUDITING PROCEDURES
o
Apparatus
I
|Acceptance
|limits
I
(Frequency and method
|of measurement
I Action if
(requirements
|are not met
Performance
audit of
analytical
phase
(Measured
(relative error of
(audit samples
(less than 15%
| (or other stated
(value) for both
|samples
1
I Frequency; Once
|during every
|enforcement source
(test" '"
(Method; Measure
|audit samples and
|compare results to
|true values
|Review
(operating
|technique and
|repeat audit
Volumetric
sampling
(Measured pretest
(volume within
|± 10% of the
(audit volume
I
I Frequency; Once
|during every
|enforcement source
(test'
(Method; -^Tleasure
|reference volume and
|compare with true
|volume
I
|Review
(operating
|techniques
O
Data
processing
errors
(Original and
|checked
|calculations
(agree within
(round-off error
I
(Frequency; Once
|during every
(enforcement test"
(Method; Independent
|calculations
|starting with
|recorded data
|Check and
|correct 'all data
|for the audit
(period
|represented
(by the
|sampled data
I ______
Systems
audit-
observance
of
technique
(Operational
|technique as
(described in
(this section of
(the Handbook
I
I Frequency; Once
|during every
|enforcement source
(source test" until
|experience gained,
|then every third
(test
(Method; Observation
(of techniques
(assisted by audit
(checklist (Fig. 4.2)
(Explain to team
(their deviations
(from recommended
|techniques and
(note on Fig. 4.2
I
•As defined here, a source test, for enforcement comprises a series of three runs at one source. Source test
for purposes other than enforcement may be audited at the frequency determined by the applicabl-s group.
O
-------
Section No. 3.19.9
Date September 3, 1992
Page 1
RECOMMENDED STANDARDS FOR ESTABLISHING TRACEABILITY
To achieve data of desired quality, two essential considerations are
necessary: (1) the measurement process must be in a state of statistical control at the
time of the measurement; and (2) the systematic errors, when combined with the random
variation (errors or measurement), must result in an acceptable uncertainty. Evidence
of quality data results from performing QC checks and independent audits of the
measurement process, documenting these data, and using materials, instruments, and
measurement procedures that can be traced to an appropriate standard of reference.
Data must be routinely obtained by repeatedly measuring standard reference
samples (primary, secondary, and/or working standards) and by establishing a condition
of process control. The working calibration standards must be traceable to standards
of higher accuracy by using a control sample or by purchasing working calibration
standards that are NIST-traceable.
Performance audit samples are not required for determining compliance;
however, an NIST control sample is recommended (as discussed in Section 3.19.8) . A
control sample is also recommended as an independent check on the measurement process
when the method is performed for other purposes. This procedure makes all the
compliance determination samples traceable to an NIST standard.
/I1-
li!
-------
o
o
o
-------
Section No. 3.19.10
Date September 21, 1992
Page 1
10.0 REFERENCE METHODS: METHOD 101A-DETERMINATION OF PARTICIPATE AND GASEOUS
MERCURY EMISSIONS FROM STATIONARY SOURCES
This method is similar to Method 101, except acidic potassium permanganate
solution is used instead of acidic iodine monochloride for sample collection.
1.0 APPLICABILITY AND PRINCIPLE
1 -1 Applicability
This method applies to determining particulate and gaseous Hg emissions from
sewage sludge incinerators and other sources, as specified in the regulations.
1-2 Principle
Particulate and gaseous Hg emissions are withdrawn isokinetically from the
source and collected in acidic potassium permanganate (KMnO<) solution. The Hg
collected (in mercuric form) is reduced to elemental Hg, which is then aerated from the
solution into an optical cell and measured by afc"3mic absorption spectrophotometry.
2.0 RANGE AND SENSITIVITY
2.1 Range
After initial dilution, the range of this method is 20 to 800 ng Hg/mL. The
upper limit can be extended by further dilution of the sample.
2.2 Sensitivity
The sensitivity of the method depends on the recorder/spectrophotometer
combination selected.
3.0 INTERFERING AGENTS
3.1 Sampling
Excessive oxidizable matter in the stack-gas prematurely depletes the KMnO<
solution and, thereby, prevents further collection of Hg. •
This section represents Method 101A and referenced procedures from Method 101. Text
from Method 101 is shown in bold italics.
-------
Section No. 3.19.10
Date September 21, 1992
Page 2
3.2 Analysis
o
Condensation of water vapor on the optical cell windows causes positive
interference.
4.0 PRECISION
Based on eight paired-train tests, the within-laboratory standard deviation
was estimated to be 4.8 jag/mL in the concentration range of 50 to 130 Jlg/m1.
5.0 APPARATUS
5.1 Sampling Train and Sample__Recovery
Same as in Method 101, Sections 5.1 and 5.2, respectively, except for the
following variations:
5.1.1 Probe Nozzle, Pitot Tube, Differential Pressure Gauge, Metering System,
Barometer, and Gas Density Determination Equipment—Same as in Method 5, Sections 2.1.1,
2.1.3, 2.1.4, 2.1.8, 2.1.9, and 2.1.10, respectively.
5.1.1 Probe Liner—Borosilicato or quartz glass tubing. Tenters may ueo a hooting
ay-atom capable of maintaining a gas tojuperaturo of 120 ± 14 °C (248 ± 25 °F) at the
probe exit during campling to prevent orator condemnation. (Notes Do not ueo metal
probo liners.)
If a filter is used ahead of the impingers, testers must use the prob
heating system to minimize the condensation of gaseous Hg. If a filter is used ahead
of the impingers, testers must use the probe heating system to minimize the
condensation of gaseous Hg.
O
5.1.2 Filter Holder (Optional)—The holder should be composed of borosilicate glass
with a rigid stainless-steel wire-screen filter support (do not use glass frit
supports) and a silicone rubber or Teflon gasket, designed to provide a positive seal
against leakage from outside or around the filter. The filter holder must be equipped
with a filter heating system capable of maintaining a temperature around the filter
holder of 120 ± 15 °C (248 ± 25 °F) during sampling to minimize both water and gaseous
Hg condensation. Testers may use a filter in cases where the stream contains large
quantities of particulate matter.
5.1.3 Impingers-Four Groonburg-Sndtb impingers are required. They should be
connected in series with leak-free ground glass fittings or any similar leak-free,
noncontominating fittings. For the firat, third, and fourth impingersf tenters may use
impingors that are modified by replacing the tip with a 13-nxn ID (0.5-in.) glass tube
extending to 13 mm (0.5 in.) from the bottom of the flask.
O
/;
-------
Section No. 3.19.10
Date September 21, 19&2
Page 3
5.2 Sample Recovery . • ... ... .
The following items are needed for sample recovery:
5.2.1 Glass Sample Bottles-The bottles should be leakless, with Teflon-lined cape,
1000 and 100 ml/.
5.2.2 Graduated Cylinder—A 250-mL graduated cylinder is required.
5.2.3 Funnel and .Rubber Policeman—These items aid in transferring silica gel to the
container; they are not necessary if the silica gel 10 weighed in the field.
5.2.4 Funnel-The funnel should be glass/ it aids in sample recovery.
5.2 Analysis
Same as in Method 101, Sections 5.3 and 5.4, respectively, except as follows:
5.2.1 Volumetric Pipets-Pipets must be Class A, 1, 2, 3, 4, 5, 10, and 20 mL.
5.2.2 Graduated Cylinder—A 25-mL graduated cylinder is required.
.2.3 Steam Bath-Same as Method 101.
5.3 Sample Preparation and Analysis
The following equipment is needed for sample preparation and analysis:
5.3.1 Atomic Absorption Spectrophotometer—Any atomic absorption unit is suitable,
provided it has an open sample presentation area in which to mount the optical cell.
Testers should follow the instrument settings recommended by the manufacturer. Instru-
ments designed specifically for measuring Hg using the cold-vapor technique are commer-
cially available and may be substituted for the atomic absorption Spectrophotometer.
5.3.2 Optical Cell—The optical cell should be cylindrical, with quartz end wlndovs
and the dimensions shown in Figure 101A-2. Wind the cell with approximately 2 m of
24-gauge nichrome heating wire and wrap with fiberglass insulation tape, or equivalent;
do not let the wires touch one another. As an alternative to the heating wire, tester-
may use a heat lamp mounted above the cell or a moisture trap installed upstream of ths
cell.
5.3.3 Aeration Cell-Tie coll must be constructed according to the specifications
in Figure 101A-3. Do not use a glass frit as a substitute for the blown glass bubbler
tip shown in Figure 101A-3. Aeration cells, available with commercial ccld-vapcr
instrumentation, may be used as an alternate apparatus.
5.3.41 Recorder—The recorder must be matched to output of the Spectrophotometer
escribed in Section 5.3.1.
V
5.5.5 Variable Transformer-The transformer is necessary for varying the voltage en
the optical cell from 0 to 40 volts.
5.3.6 Hood-A hood is required for venting the optical cell exhaust.
-------
Section No. 3.19.10
Date September 21, 1992
Page 4
5.3.7 Flow Metering Valve-Same as Method 101.
o
5.3.8 Flow Meter—A rotometor, or equivalent, la required that IB capable of
measuring a gao flow of 1.5 L/mln.
5.3.9 Aeration Gas Cylinder—The cylinder must contain nitrogen or dry, Hg-free air
and must be equipped with a single-stage regulator. As an alternative, aeration may
be provided by a peristaltic metering pump. If a commercial cold-vapor instrument is
used, follow the manufacturer's recommendations. .
5.3.10 Tubing-Tho tubing ia require for making connections. Deo glass tubing
(ungreaaed ba.ll* and socket-connections arc recommended) for all connections between
the solution cell and the optical cell; do not UBO Tygon tubing, other types of
flexible tubing, or metal tubing as eubntltuten. Teeters may use Teflon, stool, or
copper tubing between the nitrogen tank and the flow meter valve (Section 5.3.7), and
Tygon, gum, or rubber tubing between the flow meter valve and the aeration cell.
5.3.11 Flow Rate Calibration Equipment—This equipment consists of a bubble flow
meter or a wet-test meter for measuring a gas flow rate of 1.5 ± 0.1 L/mln.
5.3.12 Volumetric Flasks—Those flasks must be Class A, with ponnyhoad standard taper
stoppers; the required sizes are 100, 250, 500, and 1000 mL.
5.3.13 Volumetric Pipets-Thoso plpots must bo Class A; the required sizes are 1, 2,
3, 4, 5, 10, and 20 mL.
5.3.14 Graduated Cylinder—A 25-mL cylinder is required.
5.3.15 Magnetic Stirrer—A general-purpose laboratory-type stirring bar is required.
5.3.16 Magnetic Stirring Bar—A Teflon-coated stirring bar is required.
5.3.17 Balance-A balance capable of weighing to ± 0.5 g is required.
5.3.18 Steam Bath
O
o
-------
Section No. 3.19.10
Date September 21, 1992
Page 5
5.4 Alternative Analytical Apparatus
Alternative systems are allowable as long as they meet the following
criteria:
5.4.1 The system must generate a linear calibration curve and two consecutive
samples of the same aliquot size and concentration must agree within 3% of their
average.
5.4.2 The system must allow for recovery of a minimum of 95% of the spike when an
aliquot of a source sample is spiked with a known concentration of Hg (II) compound.
5.4.3 The reducing agent should be added after the aeration cell is closed.
5.4.4 The aeration bottle bubbler should not contain a frit.
5.4.5 Any Tygon tubing used should be as short as possible and should be
conditioned prior to use until blanks and standards yield linear and reproducible
results.
5.4.6 If manual stirring is performed before aeration, the aeration cell should be
closed during the process.
5.4.7 A drying tube should not be used unless it is conditioned following the
procedure for the Tygon tubing, above.
6.0 REAGENTS
Use ACS reagent-grade chemicals or equivalent, unless otherwise specified.
6 .1 Sampling and Recovery
The following reagents are used in sampling and recovery:
6.1.1 WaCer-Deionized distilled, meeting ASTM specifications for Type I Reagent
Water—ASTM Test Method D'1193-77. If high concentrations of organic matter are not
expected to be present, users may eliminate the KMnO4 test for oxidizable organic
matter. Use this water in all dilutions and solution preparations.
6.1.2 Nitric Acid (HNO,), 50% (v/v)-Hix equal volumes of concentrated HN03 and
water, being careful to add the acid to the water slowly.
6.1.3 Silica Gel-Indicating type, 6- to 16-mesh. If previously used, dry at 175
°C (350 °F) for 2 h. Testers may use new silica gel as received.
6.1.4 Filter (Optional;-Glass fiber filter, without organic binder, exhibiting at
least 99.95% efficiency on 0.3-nm dioctyl phthalate smoke particles. Testers may use
the filter in cases where the gas stream contains large quantities of particulate
matter, but they should analyze blank filters for Hg content. <,
6.1.5 Sulfuric Acid (H:SO.), 10% (v/v)-Add and mix 100 mL of concentrated H;SO. to
900 mL of water.
-------
Section No. 3.19.10
Date September 21, 1992
Page 6
6.1.6 Absorbing Solution, 4% KMnO, (WvJ-Prepare fresh daily. Dissolve
KMnO< in sufficient 10% H;SO, to make 1 L. Prepare and store in glass bottles to
prevent degradation. :
6.1.7 Hydrochloric Acid—Trace metals grade HCl is recommended. If other grades are
used, the Hg level must be less than 3 ng/mL Hg.
6.1.8 Hydrochloric Acid, 8 ^-Dilute 67 mL of concentrated HCl to 100 mL with water
(slowly add the HCl to the water).
6.2 Analysis
The reagents needed for analysis are listed below:
6.2.1 Tin (II) Solution-Prepare fresh daily and keep sealed when not being used.
Completely dissolve 20 g of tin (II) chloride [or 25 g of tin (II) sulfate] crystals
(Baker Analyzed reagent grade or any other brand that will give a clear solution) in
25 mL of concentrated HCl. Dilute to 250 mL with water. Do not substitute HNO3, H;SO,.,
or other strong acids for the HCl. . .
6.2.2 Sodium Chloride-Hydroxylamine Solution—Dissolve 12 g of sodium chloride and
12 g of hydroxylamine sulfate (or 12 g of hydroxylamine hydrochloride) in water and
dilute to 100 mL.
6.2.3 Hydrochloric Acid, 8 J\HDilute 67 mL of concentrated HCl to 100 mL with
(slowly add the HCl to the water).
6.2.4 Nitric Acid, 15% (v/v)-Slouly add 15 mL of concentrated HN03 to 100 mL of
water.
6.2.5 Mercury StocJc Solution, 1 mg Hg/mL-Prepare and store all Hg standard
solutions in borosilicate glass containers. Completely dissolve 0.1354 g of Hg (II)
chloride in 75 mL of water. Add 10 mL of concentrated HNO3 and adjust the volume to
exactly 100 mL with water. Mix thoroughly. This solution is stable for at least 1
month.
6.2.6 Intermediate Hg Standard Solution, 10 jig/mL-Prepare fresh weekly. Pipet 5.0
mL of the Hg stock solution (Section 6.2.5) into a 500-mL volumetric flask, and add 20
mL of 15% HNOj solution. Adjust the volume to exactly 500 mL with water. Thoroughly
mix the solution.
6.2.7 Working Hg Standard Solution, 200 ng Hg/mL-Prepare fresh daily. Pipet 5.0
mL from the Intermediate Hg Standard Solution (Section 6.2.6) into a 250-mL volumetric
flask. Add 5 mL of 4% KMnO, absorbing solution and 5 mL of 15% HNO,. Adjust che volume
to exactly 250 mL with water. Mix thoroughly.
6.2.8 Potassium Permanganate, 5% fw/vJ-Dissolve 5 g of KMnO< in water and dilute
to 100 mL.
6.2.9 Filter—Use a Whatman 40, or equivalent.
7 .0 PROCEDURE
o
-------
Section No. 3.19.10
Date September 21, 1992
Page 7
7.1.
Sampling
The sampling procedure is the same as in Method 101, except for changes
associated with using KMnO4 instead of IC1 absorbing solution and the possible use of
a filter. Because of the complexity of this method, teeters should be trained and
experienced with all procedures to ensure reliable results. Because the amount of Kg
collected generally is small, the method must be applied carefully to prevent sample
contamination or loss.
7.1.1 Pretest Preparation—Follow the general procedure given in Method 5, Section
4.1.1, but omit the directions on the filter.
7.1.2 Preliminary Determinations—The preliminary determinations are the same as
those given in Method 101, Section 7.1.2, except for the absorbing solution depletion
sign. In this method, high-oxidizable organic matter content may make it impossible
to sample for the desired minimum time. This problem is indicated by the complete
bleaching of the purple color of the KMnO< solution. In these cases, testers may
divide the sample run into two or more subruns to ensure that the absorbing solution
will not be depleted. In cases where excess water condensation is encountered, collect
two runs to make one sample. H~
7.1.2
Sampling Train Preparation
7.1.2.1 Sampling train preparation is the same as that given in Method 101, Section
7.1.3, except for the cleaning of the glassware (probe, filter holder, if used,
impingers, and connectors) and for the charging of the first three impingers. In this
method, clean all the glass components by rinsing with 50% HN03/ tap water, 8 N HCl,
tap water, and finally deionized distilled water. Then place 50 mL of 4% KMnO< in the
first impinger and 100 mL in each of the second and third impingers.
7.1.2.2 If a filter is used, place it with the filter holder with a pair of tweezers.
Be sure to center the filter, and place the gasket in the proper position to prevent
the sample gas stream from by-passing the filter. Visually check the filter for damage
after assembly is completed. Be sure to set the filter heating system at the desired
operating temperature after the sampling train has been assembled.
7.1.2.1 Follow the general procedure given in Method 5, Section 4.1.2, except as
follows: Select a nozzle size based on the range of velocity heads to ensure that it
is not necessary to change the nozzle size to maintain isokinetic sampling rates below
28 L/min (1.0 cfm).
7.1.2.2 Highly oxidizable organic content may make it impossible to sample for the
desired minimum time. This problem is indicated by the complete bleaching of the
purple color of the KMnO4 solution. If the purple color is expended in the last
(third) KMnOt impinger, the sample run is unacceptable and another run shall be
conducted. In these cases, testers may divide the sample run into two or more subruns
to ensure that the absorbing solution will not be depleted or a fourth impinger
containing 100 mL of KMnOt may be used. In cases where excess water condensation is
encountered, collect two runs to make one sample or add extra empty impinger before the
first impinger containing KMnO4 solution.
7.1.3 All the glass components should been cleaned in the laboratory (a hood is
recommended) by soaking with 50% HNO, for 1 h and then by rinsing with tap water, B N
-------
Section No. 3.19.10
Date September 21, 1952
Page 8 ......
• •<-.•••:•' •. .. • •-.(-. • . .-:••••.
HC1, tap water, and finally deionizod distilled water. After cleaning, openings shou
bo covered to prevent contamination.
7.1.3.1 Place 50 mL of 4% KMnO< in the first impinger and 100 tnL in each of the
second and third impingers. Take care to prevent the absorbing solution frc.-n
contacting any greased surfaces. Place approximately 200 g of preweighed silica gel
in the fourth impinger. Testers may use more silica gel, but they should be careful
to ensure that it is not entrained and carried out from the impinger during sampling.
Place the silica gel container in a clean place, for later use in the sample recovery.
Alternatively, determine and record the weight of. the silica gel plus impinger to the
nearest 0.5 g. (Note: Contact with KMnO4 should be avoided.)
7.1.3.2 If a filter is used, place it in the filter holder with a pair of tweezers.
Be sure to center the filter, and place the gasket in the proper position to prevent
the sample gas stream from by-passing the filter. Check the filter for tears after
assembly is completed. Be sure to set the filter heating system at the desired
operating temperature after the sampling train has been assembled.
7.1.3.3 Install tho selected nozzle using a Viton A O-ring when Btack temperatures
are less than 260 °C (500 °F). UBO a fiborgldss string gasket if temperatures are
higher. Other connecting systems uaing either 316\stainloss-stool or Teflon ferrules
may bo used. Mark the probe with heat-resistant tape or by some other method to denote
tho proper distance into the stack or duct for each sampling point. Assemble the train
an shown in Figure 101A-1,'using (if necessary) a very light coat of silicone gre&s~*<.
on all ground glass Joints. Grease only the outer portion to avoid contamination^ 1
tho oiliconc grease. ^—^
Note: An empty impinger may be inserted between the.filter and firsc impinc--£r
containing KMnO< to remove excess moisture from the sample stream.
7.1.3.4 After tho sampling train has been assembled, turn on and set the probe, if
applicable, at tho desired operating temperature. Allow time for the temperatures to
stabilize. Place crushed ice around tho iopingors.
7.1.4 Leak Check Procedures-Follow the leak chock procedures outlined in Hethod 5,
Sections A.I.A.I, 4.1.4.2, and 4.1.4.3.
7.1.3 Sampling Train Operation—In addition to the procedure given in Method.3.01,
Section 7.1.5, maintain a temperature around tho filter (if applicable) of-,120 * 14 °C
(248 ± 25 °F).
7.1.5 Mercury Train Operation-Follow the general procedure given in Method 5,
Section 4.1.5, maintain a temperature around the filter (if applicable) of 120 * 14 .°C
(248 ± 25 °F) . For each run, record tho data required on a data sheet, such as the one
shown in Figure 101A-4.
7.1.6 Calculating Percent of Isokinetic Sampling—Same as in' Method 5, Section
4.1.6.
1.2 Sample Recovery
;D
Begin proper cleanup procedure as soon as the probe is removed from the s
at the end of the sampling period. Allow the probe to cool. When it can be handled
safely, wipe off any external particulate matter near the nozzle tip and place a cap
-------
Section No. 3.19.10
Date September 21, 1951
Page 9
over it. Do not cap the probe tip tightly while the sampling train is cooling because
the resultant vacuum would draw liquid from the impingers. Before moving the sample
train to the cleanup site, remove the probe from the train, wipe off the silicons
grease, and cap the open outlet of the probe. Be careful not to lose any condensate
that might be present. Wipe the silicone grease from the impinger. Use either
ground-glass stoppers, plastic caps, or serum caps to close these openings. Transfer
the probe, impinger assembly, and (if applicable) filter assembly to an area that is
clean, protected from the wind, and free of Hg contamination.
The ambient air in laboratories located in the immediate vicinity of Hg-using
facilities is not normally free of Hg contamination. Inspect the train before and
during assembly and note any abnormal conditions. Treat the sample as follows:
7.2.1 Container Wo. 1 (Impinger, Probe, and Filter Holder) and, if Applicable, No.
1A (HC1 Rinse)
7.2.1.1 Using a graduated cylinder, measure the liquid in the first three impingers
to within 1 mL. Record the volume of liquid present (see Figure 5-3 of Method 5 in 41
CFR Part 60). This information is needed to calculate the moisture content of the
effluent gas. (Use only graduated cylinder and glass storage bottles that have beer.
precleaned as described in Section 7.1.2) Pla"ce the contents of the first three
impingers into a 1000-mL glass sample bottle. Note: If a filter is used, remove the
filter from its holder, as outlined under Container No. 3 below.
7.2.1.2 Taking care that dust on the outside of the probe or other exterior surfaces
does not get into the sample, quantitatively recover the Hg (and any condensate) fron
the probe nozzle, probe fitting, probe liner, and front half of the filter holder (if
applicable) and impingers as follows: Rinse these components with a total of 400 r.L-
of fresh 4% KMnO^ solution, carefully ensuring removal of all loose particulate matter
from the impingers. Add all washings to the 1000-mL glass sample bottle. Remove ar.y
residual brown deposits on the glassware following the permanganate rinse with
approximately 100 mL of water, carefully assuring removal of all loose particulate
matter from the impingers, and add this rinse to Container No. 1. If no visible
deposits remain after this water rinse, do not rinse with 8 N HC1. However, if
deposits do remain on the glassware after the water rinse, wash impinger walls ar.d
stems with the same 25 mL of 8 N HC1 and place the wash in a separate container labeled
Container No. 1A. Use the following procedure: Place 200 mL of water in a sample
container labeled Container No. 1A. Use only a total of 25 mL of 8 N HCl to rinse ell
impingers. Wash the impinger walls and stem with the HCl by turning and shaking the
impinger so that the HCl contacts all inside surfaces. While stirring, pour the KC1
wash carefully into Container No. 1A. The separate container is used for safety
reasons.
7.2.1.3 After all washings have been collected in the sample container, tighten the
lid to prevent leakage during shipment to the laboratory. Mark the height of the fluid
level to help determine whether leakage occurs during transport. Label the container
to identify its contents clearly.
7.2.2 Container Wo. 2 (Silica GeJ^-Note the color of the indicating silica gel tc
determine whether it has been completely spent and make a notation of its ccnditifen..
Transfer the silica gel from its impinger to its original container and seal the
container. A funnel may be used to pour the silica gel, and a rubber policeman may be
used to remove the silica gel from the impinger. It is not necessary to remove the
small amount of particles that may adhere to the impinger wall and are difficult t:
-------
Section No. 3.19.10
. • Date September' '21, 1992
. . . Page 10 -
remove. Because the weight gain is to be used for moisture calculations, do not use
any water or other liquids to transfer the silica gel. If a balance is available in
the field, weigh the spent silica gel (or silica gel plus impinger) to the nearest 0.5
g and record this weight. .
7.2.3 Container No. 3 (Filter)—If a filter was used, carefully remove it from the
filter holder, place it in a 100-mL glass sample bottle, and add 20 to 40 mL of 4%
KMnO<. If it is necessary to fold the filter, be sure that the particulate cake is
inside the fold. Carefully transfer to the 150-mL sample bottle any particulate matter
and filter fibers that adhere to the filter holder gasket by using a dry Nylon bristle
brush and a sharp-edged blade. Seal the container. Label the container to identify
its contents clearly. Mark the height of the fluid level to help determine whether
leakage occurs during transport.
7.2.4 Container No. 4 (Filter Blank)—It a filter was used, treat an unused filter
from the same filter lot used for sampling in the same manner as Container No. 3.
7.2.5 Container No. 5 (Absorbing Solution Blank)—For a blank, place 650 mL of 4 %
KMnO., absorbing solution in a 1000-mL sample bottle. Seal the container.
.-r~
7.2.6 Container No. 6 (HCl Rinse Blank)—For a blank, place 200 mL of water in a
1000-mL sample bottle. While stirring, add 25 mL of 8 N HCl. Seal the container.
Only one blank sample per 3 runs is required.
7 .3 Sample Preparation
Check the liquid level in each container to see if liquid was lost during
transport. If a noticeable amount of leakage occurred, either void the sample or use
methods subject to the approval of the Administrator to account for the losses. Then
follow the procedures below:
7.3.1 Containers No. 3 and No. 4 (Filter and Filter Blank)-I£ a filter was used,
place the contents, including the filter, of Containers No. 3 and No. 4 in separate
250-mL beakers. Heat the beakers on a steam bath until most of the liquid has
evaporated. Do not take to dryness. Add 20 mL of concentrated HNO3 to the beakers,
cover them with a watch glass, and heat on a hot plate at 70 °C for 2 h. Remove from
the hot plate. Filter the solution from the digestion of the contents of Container No.
3 through Whatman 40 filter paper and save the filtrate for addition to the Container
No. 1 filtrate, as described below. Discard the filter. Filter the solution from the
digestion of the contents of Container No. 4 through Whatman 40 filter paper and save
the filtrate for addition to the Container No. 5 filtrate, as described in Section
7.3.2 below. Discard the filter.
7.3.2 Container No. 1 (Impingers, Probe, and Filter Holder) and, if Applicable, No.
1A (HCl Rinse)—Filter the contents of Container No. 1 through Whatman 40 filter paper
into a 1-L volumetric flask to remove the brown MnO_ precipitate. Save the filter.
'Add the sample filtrate from Container No. 3 to the 1-L volumetric flask and dilute to
volume with water. If the combined filtrates are greater than 1000 mL, determine the
volume to the nearest mL and make the appropriate corrections for blank subtractions.
Mix thoroughly.
Mark the filtrate as Analysis Sample No. A.I and analyze for Hg within 48 h
after completing the filtration step. Place the saved filter, which was used to remove
the brown MnO: precipitate, into a container of appropriate size. Add 25 mL of £ N HCl
o
o
o
-------
Section No. 3.19.10
Date September 21, 1992
Page 11
to the filter and allow the filter, with its brown residue, to digest for a minimum of
24 h at room temperature. Filter the contents of Container No. 1A through a Whatman
40 filter paper into a 500-mL volumetric flask. Then filter the digestion of the brown
MnO; precipitate from Container No. 1 and the Whatman paper filter into the 500-mL
volumetric flask. Dilute to volume with water. Mark this 500-mL dilute solution as
'Analysis Sample No. HCl A.2 and analyze for Hg. Discard the filters.
7.3.3 Containers No. 5 (Absorbing Solution Blank) and No. 6 (HCl Rinse Blank)—Treat
Container No. 5 the same as Container No. 1, described in the previous section. Add
the filter blank filtrate from Container No. 4 to the 1-L volumetric flask and dilute
to volume. Mix thoroughly. Mark this as Sample No. A.I blank and analyze for Hg
within 48 h after completing the filtration step. Digest any brown precipitate
remaining on the filter from the filtration of Container No. 5, using the procedure
described in Section 7.3.2. Filter the contents of Container No. 6 using the procedure
described in Section 7.3.2 and combine into the 500-mL volumetric flask with the
filtrate from the digested blank MNO2 precipitate. Mark this resultant 500-mL combined
dilute solution as Analysis Sample No. HCl A.2 blank. Note: When analyzing blank
samples A.I blank and HCl A.2 blank, always begin with 10-mL aliquots. This note
applies specifically to blank samples.
--*"""
7.4 Analysis .
7.4.1 Calibrate the spectrophotometer and recorder and prepare the calibration
curve as described in sections 8.1 and 8.2. Then repeat the procedure used to
establish the calibration curve with aliquots of appropriate size (1 to 10 mL) of the
samples (from sections 7.3.2 and 7.3.3) until two consecutive peak heights agree within
3% of their average value. If the 10-mL sample is below the detectable limit, use a
larger aliquot (up to 20 mL), but decrease the volume of water added to the aeration
cell accordingly to prevent the solution volume from exceeding the capacity of the
aeration bottle. If the peak maximum of a 1-mL aliquot is off scale, further dilute
the original sample to bring the Hg concentration into the calibration range of the
spectrophotometer. If the Hg content of the absorbing solution and filter blank is
below the working range of the analytical method, use zero for the blank.
7.4.2 Run a blank and standard at least after every five samples to check the
spectrophotometer calibration; recalibrate as necessary. It also is recommended that
at least one sample from each stack test be checked by the Method of Standard Additions
to confirm that matrix effects have not interfered with the analysis.
8.0 Calibration and Standards
The calibration and standards are the same as in Method 101, Section 8,
except for the following variations:
8.1 Optical Cell Heating System Calibration
Same as in Method 101, Section 8.2, but use a 25-mL graduated cylinder to add
25 mL of water to the bottle section of the aeration cell.
i,
8.2 Spectrophotometer and Recorder Calibration
8.2.1 The Hg response may be measured by either peak height or peak area. Note:
The temperature of t,he solution affects the rate at which elemental Hg is released;
-------
Section No. 3.19.10
Date September 21, 1992
Page 12
consequently, it affects the shape of the absorption curve (area) and the point of
maximum absorbance (peak height). To obtain reproducible results, all solutions must
be brought to room temperature before use.
8.2.2 Set the spectrophotometer wave length at 253.7 nm and 'make certain the
optical cell is at the minimum temperature that will prevent water condensation. Then
set the recorder scale as follows: Using a 25-mL graduated cylinder, add 25 mL of
water to the aeration cell bottle, and pipet 5 mL of the working Hg standard solution
into the aeration cell. Note: Always add the Hg-containing solution to the aeration
cell after the 25 mL of water.
8.2.3 Place a Teflon-coated stirring bar in the bottle. Add 5 mL of the 4% KMn04
to the aeration bottle and mix well. Attach the bottle section to the bubbler section
of the aeration cell. Make certain that: (1) the aeration cell exit arm stopcock
(Figure 101-3 of Method 101) is closed (so that Hg will not prematurely enter the
optical cell when the reducing agent is being added); and (2) there is no flow through
the bubbler. Add 5 mL of sodium chloride hydroxylamine in 1-mL increments until the
solution is colorless. Now add 5 mL of tin (II) solution to the aeration bottle
through the side arm and immediately stopper the side arm. Stir the solution for 15
s, turn on the recorder, open the aeration cell" exit arm stopcock, and immediately
initiate aeration with continued stirring. Determine the maximum absorbance of the
standard, and set this value to read 90% of the recorder full scale.
Before uao, clean nil glao8wa.ro, both now and unod, as follows: Brush with
soap and tap water, liberally rinse with tap water, soak for 1 h In 50% HNO3. Rinse
with dolonlzod distilled water.
6.1 Flow Calibration
Assemble the aeration system an shown in Figure 101-5. Sot the outlet
pressure on the aeration gas cylinder regulator to a minimum pressure of 500 mm Hg (10
pel), and use the flow motoring valve and a bubble flow meter or wot-tost meter to
obtain a flow rate of 1.5 ± 0.1 L/mln through the aeration cell. After the flow
calibration IB completed, remove the bubble flow noter from the system.
8.2 Optical Cell Heating System_ Calibration
Using a 25-mL Graduated cylinder/ add 25 mL of water to the bottle section
of the aeration cell and attach the bottle section to the bubbler section of the cell.
Attach the aeration coll to the optical coll; while aerating at 1.5 L/nln, determine
tho minimum variable transformer sotting necessary to prevent condensation of moisture
In the optical cell and In the connecting tubing. (This setting should not exceed 20
volts.)
8.3 Spectrophotometer and Recorder Calibration
8.3.1 The Hg response may bo measured by either peak height or peafc area. fWote:
The temperature of the solution affects the rate at which elemental Hg ID released;
conse
-------
Section No. 3.19.10
Date September 21, 1952
Page 13
Bet the recorder scale as follows.- Using a. 25-mL Graduated cylinder, add 25 mL of
water to the aeration cell bottle and pipet 5 aL of the working Hg standard solution
into the aeration cell. (Note: Always add the Hg-containing solution to the aeration
cell after the 25 mL of water.)
8.3.3 Place a Teflon-coated stirring bar in the bottle. Using a 25-mL graduated
cylinder, add 25 mL of laboratory pure water to the aeration cell bottle. Pipet 5.0
mL of the working Hg standard solution to the aeration cell. Add 5 mL of the 4% KMnO,
absorbing solution, followed by 5 mL of 15% UNO, and 5 mL of 5% KMnO< to the aeration
cell, and mix well using a swirling motion. Attach the bottle to the aerator, making
sure that: (1) the exit arm stopcock is closed, and (2) there is no aeration gas
flowing through the bubbler. Through the side arm, add 5 mL of sodium chloride
hydroxylamine solution in 1 mL-increments until the solution is colorless. Through the
side arm, add 5 mL of the Tin (II) reducing agent to the aeration cell bottle, and
immediately stopper the side arm. Stir the solution for 15 s and turn on the recorder
or integrator. Open the aeration cell exit arm stopcock and initiate the gas flow.
Determine the maximum height (absorbance) of the standard, and set this value to read
90% of the recorder full scale.
8.4 Calibration Curve -->-
8.4.1 After setting the recorder scale, repeat the procedure in Section 8.3 using
-, 1-, 2~, 3-, 4-, and 5-mL aliquots of the working standard solution (final amount
f Kg In the aeration cell Is 0, 200, 400, 600, BOO, and 1000 ng, respectively).
Repeat this procedure on each aliquot size until two consecutive peaks agree within 3%
of their average value. (Note: To prevent Hg carryover from one sample to another, do
not close the aeration cell from the optical cell until the recorder pen has returned
to the baseline.)
8.4.2 It should not be necessary to disconnect the aeration gas inlet line from the
aeration cell when changing samples. After separating the bottle and .bubbler sections
of the aeration cell, place the bubbler section into a 600-mL beaker containing
approximately 400 mL of water. Rinse the bottle section of the aeration cell with a
stream of water to remove all traces of the tin (II) reducing agent. Also, to prevent
the loss of Hg before aeration, remove all traces of the reducing agent between samples
by washing with water. It will be necessary, however, to wash the aeration cell parts
with concentrated HCl if any of the following conditions occur: (1) a white film
appears on any inside surface of the aeration cell; (2) the calibration curve changes
suddenly; or (3) the replicate samples do not yield reproducible results.
8.4.3 subtract the average peak height (or peak area) of the blank (0-aL
aliquot)-which should be less than 2% of" recorder full scale-from the averaged peak
heights of the 1-, 2-, 3-, 4-, and 5-mL aliquot standards. If the blank absorbance is
greater than 2% of full-scale, the probable cause is Hg contamination of a reagent or
carry-over of Hg from a previous sample. Plot the corrected peak height of each
standard solution versus the corresponding final total Hg weight in the aeration cell
(in ng), and draw the best-fit straight line. This line should either pass through the
>rigin or pass through a point no further from the origin than * 2 % of the recorder
ull scale. If the line does not pass through or very near to the origin, check for
"nonlinearity of the curve and for incorrectly prepared standards.
9.0 Calculations
-------
Section No. 3.19.10
Date September 21, 1552
Page 14
9.1 Dry-Gas Volume, Volume of Water Vapor and Moisture Content, Stack-S=!
Velocity, Isokinetic Variation and Acceptable Results, and Determination of Compliance
Same as in Method 101, Sections 9.1, 9.2, 9.3, 9.6, and 9.7, respectively,
but use data obtained from this test.
9.1 Dry-Gas Volume
Using the data from this toot/ calculate Vflttll), the dry-gas sample volume at
standard conditions (corrected for leakage, if necessary) as outlined in Section 6.3
of Method 5. '
9.2 Volume of Water Vapor and Moisture Content
Using the data obtained from this tost, calculate the volume of water vapor
Vul,ca, and the moisture content Bw, of the stack-gas. Use equations 5-2 and 5-3 of
Method 5.
9.3 Stack-Gas Velocity
-•*"""
Using the data from this teat and Equation 2-9 of Method 2, calculate the
average stack-gas velocity v,.
9.4 Isokinetic Variation and Acceptable Kesults ' • ; "
Same as in Method 5, Sections 6.11 and 6.12, respectively.
9.5 Determining Compliance
Each performance test consiots of three repetitions of the applicable tesc
.method. For the purpose of determining compliance with an applicable national emission
standard, use the average of the results of all repetitions.
9.2 Total Mercury
For each source sample, correct the average maximum absorbance of the tvrc
consecutive samples whose peak heights agreed within 3% of their average for the
contribution of the blank. Then calculate the total Kg content in |Jg in each sample.
Correct for any dilutions made to bring the sample into the working range of th<=
spectrophotometer.
O
O
-------
Section -No; 3.15.10
Date September 21, 1952
Page 15
[C(HC1 Jb-Z *)•„;, DFt:.K]
~~~
Equation 101A-1
where:
m(HCl)H,
C(HC1),
C(HC1 blk)H,
DF
DFM,
V(IH.:
10"
S
Total blank corrected ng of Hg in HC1 rinse and HC1
digestate of filter sample. ...
Total ng of Hg analyzed ;in.^the aliquot from the BOO-iri
Analysis Sample No. HCLA.2.,
Total ng of Hg analyzed .in aliquot of the 500-rnL Anal-
ysis Sample No. HC1_A.2 blank.
Dilution factor for the HC1 digested Hg-ccntaining
solution, Analysis Sample No. HC1 A.2. This dilution
factor (DF) applies only to the intermediate dilution
steps because the original sample volume (V.),.-: of HCi
A.2 has been factored out in the equation, along with
the sample aliquot, (S) . In Equation 6.9, the sample
aliquot, S, is introduced directly into the aeration
cell for analysis according to the procedure outlined in
Section 3.19.5.3.4. A dilution factor is required only
if it is necessary to bring the sample into the analyti-
cal instrument's calibration range. If no dilution is
necessary, then DF equals 1.0.
Dilution factor for the Analysis Sample No. HCI A. 2
blank. (Note: Normal dilution factor calculations apply
here.)
Solution volume of original sample, 500 mL for samples
diluted as described in Section 7.3.1.
Conversion factor, Jlg/ng.
Aliquot volume of sample added to aeration cell, mL.
Aliquot volume of blank added to aeration cell, ~iL.
Note: The maximum allowable blank subtraction fcr the HCI is the lesser cf
the two following values: (1) the actual blank measured value (Analysis Sample I'io. HCI
.2 blank); or (2) 5% of the Hg content in the combined HCI rinse and digesced sampi*
'(Analysis Sample No. HCI A.2).
where:
-------
Section No. 3.19.10
Date September 21, 199
Page 16
o
DF
(C(fltrblk) DFblk
m(fltr).
C(fltr)
C(fltr blk).
V
f ibH.i
10~? Equation 101A-2
Jblk
= Total blank corrected Hg of Hg in KMNO4 filtrate and HNC;
digestion of filter sample.
= Total ng of Hg in aliquot of KMNO4 filtrate and HNC-.
digestion of filter analyzed (aliquot of Analysis Sample
No. A.1).
= Total ng of Hg in aliquot of KMNO< blank and HNO3 diges-
tion of blank filter analyzed (aliquot of Analysis
Sample No. A.I blank).
= Solution volume of Original sample, normally 1000 mL fcr
samples diluted as described in Section "7.3.2.
= Solution volume of blank sample, 1000 mL for samples
diluted as described in Section 7.3.2
Note: The maximum allowable blank subtraction for the HC1 is the lesser cf
the two following values: (1) the actual blank measured value (Analysis Sample No. A.I
blank); or (2) 5% of the Hg content in the filtrate (Analysis Sample No. A.I).
O
= m(HCl}H. + m(fltr)H
Equation 101A-3
where:
9.3
m(fltr)H,
= Total blank corrected Hg content in each sample, \ig.
= Total blank corrected ug of Hg in HCl rinse and HCl digestate
of filter sample.
= Total blank corrected Hg of Hg in KMNO,. filtrate ar.c HNC:.
digestion of filter sample.
Mercury Emission Rate
Calculate the Hg emission rate R in g/day for continuous operations using
Equation 101A-4. For cyclic operations, use only the time per day each stack is ir.
operation. The total Hg emission rate from a source will be the summation of results
from all stacks. <• .
v;here:
O
Total blank corrected Hg content in each sample, ug.
-------
A.
86,400
10'"
T.
Ps
K
Section No. 3.19.10
Date September 21, 1992
Page 17
R = K
Ha Vs Ar (86,400 X
V + V I ( T IP )
"mlred! VK,£iJ\ -I v •'-' cf'
Average stack-gas velocity, m/sec (fps).
Stack cross-sectional area, nr (ft*).
Conversion factor, s/day.
Conversion factor,
Equation 101A-4
Dry-gas sample volume at standard conditions, corrected for
leakage (if any), nr (ftj).
Volume of water vapor at standard conditions, m3 (ft3) .
Absolute stack-gas temperature, °K (°R).
Absolute stack-gas pressure" mm Hg (in. Hg) .
0.3858 °K/mm Hg for metric units.
17.64 °R/in. Hg for English units.
-------
Section No. 3.19.10
Date September 21. 1992
Page 18
10.1 Biblioqraphy
1. Same as bibliography in Method 101.
2. Mitchell, W.J., M.R. Midgett, J.C, Suggs, and D. Albrinck. Tesz
Methods to Determine the Mercury Emissions from Sludge Incineraticn
Plants. EPA-600/4-79-058. September 1979. U.S. Environmental
Protection Agency (EPA). Research Triangle Park, NC.
3. ' Wilshire, Frank W., J.E. Knoll, T.E/ Ward, and M.R. Midgett.
Reliability Study of the U.S. EPA's Method 101A - Determination of
Particulate and Gaseous Mercury Emissions. Report No. 600/D-31/21&
AREAL 367, NTIS Ace No. PB91-233361. U.S. Environmental Protectior.
Agency (EPA). Research Triangle Park, NC.
o
o
o
-------
Section No.3.19.11
Date September 3, 1952
Page 1
11.0 REFERENCES
1. Method 101A - Determination of Particulate and Gaseous Mercury Emissions from
Sewage Sludge Incinerators. Federal Register, Volume 47, July 8, 1982, p.
24703.
2. Corrections to Method 101A-, Federal Register, Volume ,49, September 12, 1984,
p. 35768.
3. Corrections to Method 101A. Federal Register, Volume 53, September 23, 1988,
p. 36972. , .
4. Method 101 - Determination of Particulate and Gaseous Mercury Emissions from
Chlor-Alkali Plants - Air Streams. Federal Register, Volume 38, May 6, 1973,
p. 08826.
5. Amendments to Method .101. Federal Register, Volume 47, July 8, 1982, p.
24703.
6.
8.
Corrections to Method 101. Federal Register, Volume 49, September 12, 1984,
p. 35768.
Corrections to Method 101. Federal Register, Volume 53, September 23, 1988,
p. 36972. ' "' '
Wilshire, Frank W., J.E. Knoll, T.E. Ward, and M.R. Midgett. Reliability
Study of the U.S. EPA's Method 101A - Determination of Particulate .and
Gaseous Mercury Emissions. Report No. 600/D-31/219 AREAL 367, NTIS Ace No.
PE91-233361, U.S. Environmental Protection Agency, Research Triangle Park,
NC. .. " . .
9. Addendum to Specifications for Incinerator Testing at Federal Facilities.
PHS, NCAPC. December 6, 1967.
10. Determining Dust Concentration in a Gas Stream. ASME Performance Test Code
No. 27. New York, NY. 1957.
11.
12.
13.
DeVorkin, Howard, et al. Air Pollution Source Testing Manual. Air Pollution
Control District. Los Angeles, CA. November 1963. -:
Hatch, W.R., and W.I. Ott. Determination of Sub-Microgram Quantities of
Mercury by Atomic Absorption Spectrophotometry. Anal. Chem. 40:2085-87, 1968.
Mark, L.S. Mechanical Engineers' Handbook. McGraw-Hill Book Co., Inc. New
York, NY. 1951.
Martin, Robert M. Construction Details of Isokinetic Source Sampli'ng
Equipment. EPA APTD-0581, U.S. Environmental Protection Agency. Research
Triangle Park, NC. April 1971. :'y
-------
o
Section No.3.19.11
Date September 3, 1992
Page 2
15. Western Precipitation Division of Joy Manufacturing Co. Methods for
Determination of Velocity, Volume, Dust and Mist Content of Gases. Bulletin
WP-50. Los Angeles, CA. 1968.
16. Perry, J.H. Chemical Engineers' Handbook. McGraw-Hill Book Co., Inc. New
York, NY. 1960.
17. Rom, Jerome J. Maintenance, Calibration, and Operation of Isokinetic Source
Sampling Equipment. EPA APTD-0576, U.S. Environmental Protection Agency-
Research Triangle Park, NC. April 1972.
18. Shigehara, R.T., W.F. Todd, and W.S. Smith. Significance of Errors in Stack
Sampling Measurements. Stack Sampling News. 1{3):6-18, September 1973.
19. Smith, W.S., et al. Stack Gas Sampling Improved and Simplified with New
Equipment. APCA Paper No. 67-119. 1967.
20. Smith, W.S., R.T. Shigehara, and W.F. Todd. A Method of Interpreting Stack
Sampling Data. Stack Sampling News. 1(2):8-17, August 1973.
21. Specifications for Incinerator Testing at Federal Facilities. PHS, NCAPA.
1967.
22. Standard Method for Sampling Stacks for Particulate Matter. In: 1971 Annual F j
Book of ASTM Standards, Part 23. ASTM Designation D 2928-71. Philadelphia, PA ^—'
1971.
23. Vennard, J.K. Elementary Fluid Mechanics. John Wiley and Sons, Inc. New York.
1947.
24. Mitchell, W.J., and M.R. Midgett. Improved Procedure for Determining Mercury
Emissions from Mercury Cell Chlor-Alkali Plants. J. APCA. 26:674-677, July
1976.
25. Shigehara, R.T. Adjustments in the EPA Nomograph for Different Pitot Tube
Coefficients and Dry Molecular Weights. Stack Sampling News. 2:4-11, October
1974.
26. Vollaro, R.F. Recommended Procedure for Sample Traverses in Ducts Smaller
than 12 Inches in Diameter. U.S. Environmental Protection Agency, Emission
Measurement Branch. Research Triangle Park, NC. November 1976.
27. Klein, R., and C. Hach. Standard Additions: Uses and Limitation in Spectro-
photometric Measurements. Amer. Lab. 9:21, 1977.
28. Water, Atmospheric Analysis. In: Annual Book of ASTM Standards, Part 31. ASTM
Designation D 1193-74. Philadelphia, PA. 1974.
29. Mitchell, W.J., M.R. Midgett, J.C. Suggs, and D. Albrinck. Test Methods to
Determine the Mercury Emissions from Sludge Incineration Plants. EPA-600-
/4-79-058, U.S. Environmental Protection Agency. Research Triangle Park, NC.
September 1979.
O