I
m                  UNITED STATES ENVIRONMENTAL PROTECTION AGENCY
I                             OFFICE OF ENFORCEMENT
I
I
I                EPA-330 I 9-94-001
I                  SUMMARY OF NEIC LEAK DETECTION
                  AND REPAIR PROGRAM INVESTIGATIONS
1      '                       '%,
•                January 1994
I
I
I
I
I
I
I
I
|                   NATIONAL ENFORCEMENT INVESTIGATIONS CENTER
•                                 Denver, Colorado
I
I

-------
I
I
I
I
I
I
I
I
 I
 I
 I
 I
 I
 I
  I
  i
  i
  i
  i
                      CONTENTS



INTRODUCTION  	 1


NEIC LDAR PROGRAM INVESTIGATION METHODS	 3

     FILE AND DOCUMENT REVIEW	 3
     EVALUATION	 3
     COMPONENT MONITORING	 4
     DATA EVALUATION AND INVESTIGATION REPORT	 4

SUMMARY OF THE NEIC LDAR PROGRAM INVESTIGATIONS	 5



APPENDIX


Reference Method 21

-------
I
I
I
I
                     Fugitive VOC emissions are regulated by 40 CFR Parts 60 and 61.
•              Specific regulations and requirements for National Emission Standards for
I
I
I
I
•
I
I
                                         INTRODUCTION

                     The Clean Air Act requires facilities with equipment in volatile organic
                compound (VOC) service  to develop and implement a Leak Detection and
                Repair (LDAR) program to control fugitive VOC emissions.
                Hazardous Air Pollutants (NESHAP), New Source Performance Standards
                (NSPS), and additional VOC components are outlined in various subparts of
                40 CFR Parts 60 and 61. State and local agencies normally incorporate the
                federal requirements by reference into the state and local air regulations, but
                                             *•                     v •»
                may have their own LDAR requirements.

                     NESHAP and NSPS regulations  require regulated  facilities  with
                equipment operating in VOC service to implement an LDAR program designed
•              to reduce fugitive emissions. Requirements include equipment identification,
"              a monitoring program, and recordkeeping and reporting. The leak monitoring
•              requirements apply to all plant equipment containing 10% or greater VOC,
                including  pumps,  valves,  compressors,  and  flanges.   VOC  components
•              associated with storage tanks, loading/unloading areas, and transfer lines are
                also subject to monitoring requirements.

                     Facilities may choose to operate under alternative standards for valves
I              allowing a 2.0% total valve leak rate for each unit.  Component monitoring of
                valves in benzene service and valves in NSPS units can be conducted annually,
I              as outlined in 40 CFR § 61.112 and 40 CFR § 60.483.
                     NEIC conducts comprehensive LDAR program investigations, often as
               part of multi-media compliance investigations, to determine compliance with
               LDAR program requirements. The NEIC LDAR on-site inspections are divided

-------
into two parts: (1) monitoring of VOC components by NEIC personnel, and
(2) evaluating facility monitoring procedures. NEIC monitors between 2,000

EPA Reference Method 21" [Appendix]. This report provides an overview of the

NEIC LDAR program investigations.
I
I
                 and 3,000 components at each inspected facility using procedures specified in
                 EPA Reference Method 21" [Appendix]. This report provides an overview of thi
                 NEIC LDAR investigation methods and  summarizes the results of recent
I
I
I
I
I
I
I
I
I
I
I
I              	
•                  *  As specified in 40 CFR Part 60 Appendix A
                                                   2
I
I

-------
Quarterly VOC monitoring reports are reviewed and are used to select process
units for monitoring.
I
                          NEIC LDAR PROGRAM INVESTIGATION METHODS

_                    NEIC LDAR investigations include:

                      I*     Review of EPA and state files and documents
                      •     Evaluation of the facility monitoring procedures
                      I*     On-site VOC component monitoring
                      •     Data evaluation and report preparation

"              FILE AND DOCUMENT REVIEW

•                    State and federal files are reviewed for information regarding facility
                                              —                     *•_
•              operations/equipment, permits, inspections, and  any enforcement actions.


I
•              EVALUATION

•                    NEIC evaluates fugitive VOC monitoring procedures used by facilities
                or facility contractors. The evaluation includes:

I
                      *     Determination of facility processes  involving VOCs through
I                          discussions with facility personnel, including a detailed process
                            description
I                    •     Review of facility/contractor monitoring procedures,  including
                            equipment calibration
J                    •     Review of  VOC  component inventory, leak logs, repair logs,
                            emission reports
|                    •     Identification of specific components in VOC service

I

I

I

-------
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
I
I
 I
I
I
 COMPONENT MONITORING



       NEIC monitors a representative number of components in VOC service

 from process units, storage tanks, loading/unloading areas, transfer lines, and

 ancillary equipment using an organic vapor analyzer. Components monitored

 are selected using various criteria including:



    .  •    Equipment subject to NSPS and NESHAP requirements

      •    Past monitoring results  .

      •    Presence of "special" types of VOC equipment, such  as railroad

           car and truck loading racks

      •    Type of VOC in service



DATA  EVALUATION AND INVESTIGATION REPORT



     Upon  completion  of  the  on-site inspection,  NEIC evaluates the

information/data obtained.   This includes data collected during component

monitoring to determine equipment leak rates.  Leak rates determined by

NEIC are compared to leak rates reported by the facility.  An investigation

report is then prepared and provided to Regional personnel.

-------
 I
 I
 I
 I
 I
 I
 I
 I
 I
 •
 I
 I
 I
 I
 I
 I
 I
I
I
     SUMMARY OF THE NEIC LDAR PROGRAM INVESTIGATIONS



      None of the facility LDAR programs evaluated by NEIC were in full

compliance with regulatory requirements.  Although the facilities inspected

were located in various states and regions, the common deficiencies outlined

below were identified. These deficiencies have most likely resulted in greater

uncontrolled VOC emissions.



*     VOC component leak rates determined by NEIC are greater than the

      leak  rates reported by the majority of facilities for each process unit

      inspected. NEIC leak rates and facility reported leak rates for several

      process units are summarized below:
      Facility Reported Percent Leak Rate


                2.7
                1.6
                3.5
                0.73
                1.6
                4.0
                1.3
                0.72
                1.7
                4.1
                0.8
                0.8
                0.4
                4.4
NEIC Determined Percent Leak Rate


         18.7
          6.1
         15.4
          8.4
          6.7
         13.0
          6.9
          5.4
          6.3
         13.0
          6.1
          7.7
          7.2
          2.9
     Several process units monitored by NEIC were operating under the

     alternative standards for valves allowing a 2.0% valve leak rate for each

     unit.  This allows facilities to monitor valves in benzene service and

     valves in  NSPS units annually rather  than quarterly.  Facilities

     operating under this standard and monitored by NEIC exceeded the 2%

     allowable percentage of valves leaking.

-------
I
_              .  •     Inaccurate/incomplete VOC component inventories

_                     -     Many  regulated  components  are  not  identified/tagged  for
I                           inclusion into the monitoring program
I                             Equipment monitored not in VOC service (i.e. hot water, caustic,
                             and steam lines)
I
I
I
Incorrect/inconsistent monitoring procedures
I                     -     Components in VOC service are not monitored
                             Inadequate monitoring equipment calibration procedures
I                     -     Method 21 procedures not followed

•                     NEIC also identified inconsistencies within local LDAR regulations and
                 deficiencies between local and federal regulations. Conflicting requirements
•               within particular local  regulations  caused ambiguity in determining the
                 number of leaking components at some facilities.
I
                       NEIC identified other sources of fugitive VOC emissions not subject to
J               LDAE requirements. For example, gasket seals on rotary filters, not included
                 in LDAR requirements, were found to be leaking VOCs at concentrations in
|               excess of 10,000 parts per million.

I

I

I

I

-------
I
I
I              .
I
I
I
•                                APPENDIX
•                           REFERENCE METHOD 21
I
I
I
I
I
                    *»'
I
I
I
I
I

-------
1

I

I

I

I

I

I

 I

 I

 I

 I

 I

 I

 I

 I

  I

  I

  I

  I
 PI. 60, App. A, Meth. 21

 be  on a dry basis.  Calculate the pollutant
 emission rate, as follows:
                   20.9
E=C«Fd    209_
                                Eq. 20-6
where:
E = Mass emission rate  of pollutant. ng/J
   (lb/10« Btu).
  7.5.2  Calculation of Emission Rate Using
Carbon Dioxide Correction. The COj con-
centration and the pollutant concentration
may be on either a dry basis or a wet basis,
but  both concentrations  must be  on the
same basis for the calculations.  Calculate
the pollutant emission rate using Equation
20-7 or 20-8:
       E=C«FC
      E=C.FC
            100

           %CO,
            100

          %co»w
Eq. 2(1-7
Eq. 20-8
where:
Cw=Pollutant concentration measured on a
    moist sample basis, ng/sm3 (Ib/scf).
%COJw=Measured CO, concentration meas-
    ured on a moist sample basis, percent.
8. Bibliography
  1. Curtis, P. A Method for Analyzing NO.
Cylinder Gases-Specific Ion Electrode Pro-
cedure, Monograph available from Emission
Measurement Laboratory, ESED, Research
Triangle Park. NC 27711, October 1978.
  2. Sigsby, John  E..  P.  M. Black, T. A.
Bellar,  and D. L  Klosterman.  Chemilu-
minescent Method for Analysis of Nitrogen
Compounds in  Mobile Source  Emissions
(NO, NOi,  and NHa ). "Environmental  Sci-
ence  and  Technology."  7:51-54. January
1973.
  3. Shigehara.  R.T.,  R.M. Neulicht,  and
W.S. Smith. Validating Orsat Analysis Data
from  Possil Fuel-Fired   Units.  Emission
Measurement Branch,  Emission Standards
and Engineering Division,  Office  of  Air
Quality Planning and Standards, U.S. Envi-
ronmental Protection Agency, Research Tri-
angle Park. NC 27711. June 1975.
  METHOD 21—DETERMINATION or VOLATILE
        ORGANIC COMPOUNDS LEAKS
1. Applicability and Principle
          40 CFR Ch. I (7-L92 Edition)

  1.1  Applicability. This method applies to
the determination of volatile organic com-
pound (VOC) leaks from process equipment.
These sources include, but are not limited
to.  valves,  flanges and  other connections.
pumps and compressors,  pressure relief de-
vices, process  drains, open-ended valves,
pump and compressor seal system degassing
vents,  accumulator  vessel vents,  agitator
seals, and access door seals.
  1.2  Principle. A  portable instrument  is
used to detect  VOC leaks from individual
sources. The instrument detector type is not
specified,  but it must meet  the  specifica-
tions and performance criteria contained in
Section 3. A leak  definition  concentration
based on a reference compound is specified
in each applicable  regulation. This proce-
dure is intended to locate and classify leaks
only, and is not to be used as a direct meas-
ure of mass emission rates from individual
sources.
2. Definitions
  2.1  Leak Definition-Concentration.  The
local VOC concentration  at the surface of a
leak source that indicates that a VOC emis-
sion (leak) is present. The leak definition is
an instrument meter reading based on a ref-
erence compound.
  2.2  Reference Compound. The VOC spe-
cies  selected as an instrument calibration
basis for specification of  the leak definition
concentration. (For example: If a leak defi-
nition  concentration is  10,000 ppmv as
methane, then any source emission that re-
sults in a local concentration that yields a
meter reading of 10.000  on an instrument
calibrated with methane  would be classified
as a leak. In this example, the leak defini-
tion is 10,000 ppmv, and the reference com-
pound is methane.)
  2.3  Calibration Gas. The VOC compound
used to adjust the instrument  meter reading
to a known value. The calibration gas is usu-
ally the reference compound at a concentra-
tion approximately equal to the leak defini-
tion concentration.
  2.4  No  Detectable Emission.  Any  VOC
concentration at a potential leak source (ad-
justed for local VOC ambient  concentra-
tion) that is less than a value  corresponding
to the instrument readability specification
of section 3.1.1(c) indicates that a leak is not
present.
  2.5  Response Factor.  The ratio of the
known concentration of  a VOC compound
to the observed meter reading when meas-
ured using an  instrument calibrated with
the reference compound specified in the ap-
plication regulation.
  2.6  Calibration Precision. The degree of
agreement between  measurements of the
same known value, expressed as the relative
percentage of  the  average difference be-
                                       1020

-------
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
I
I
 Environmental Protection Agency

 tween the meter readings and the  known
 concentration to the known concentration.
  2.7 Response Time!  The time interval
 from a step change in VOC concentration at
 the  input of  the sampling system  to the
 time at which 90 percent of  the correspond-
 ing  final  value is reached as  displayed on
 the instrument readout meter.
 3. Apparatus
  3.1 Monitoring Instrument.
  3.1.1  Specifications.
  a.  The VOC instrument detector shall re-
 spond to the compounds being processed.
 Detector  types which may meet this  re-
 quirement include,  but are  not limited to,
 catalytic oxidation,  flame ionization, infra-
 red absorption, and photoionization.
  b.  Both the linear response range and the
 measurable range of the  instrument for
 each of the VOC to be measured, and for
 the  VOC calibration  gas that is  used for
 calibration, shall encompass the leak defini-
 tion concentration specified in the regula-
 tion. A dilution probe assembly may beJUsed
 to bring the VOC concentration within both
ranges;  however,  the  specifications  for in-
 strument  response time  and sample probe
 diameter shall still be met.
  c. The scale of the instrument meter shall
 be readable to ±2.5  percent  of the specified
 leak definition concentration when perform-
 ing a no detectable emission survey.
  d.  The instrument shall be equipped with
 an electrically driven pump to  insure that a
 sample is provided to the detector at a con-
 stant flow rate. The  nominal sample flow
 rate, as measured at the sample probe tip,
 shall be 0.10 to 3.0  liters  per minute when
 the probe is fitted with a glass wool plug  or
 filter that may be used to prevent plugging
 of the instrument.
  e.  The  instrument  shall  be intrinsically
 safe  as defined by the applicable  U.S.A.
 standards (e.g.. National  Electric  Code by
 the  National Fire Prevention Association)
 for operation in any explosive atmospheres
 that may be encountered in its use. The in-
 strument  shall, at a minimum, be intrinsi-
 cally safe for Class  1,  Division 1 conditions,
 and  Class 2. Division 1 conditions, as defined
 by the example Code. The instrument shall
 not be operated with any safety device, such
 as an exhaust flame arrester, removed.
  f. The instrument shall be equipped with
 a probe or probe extension for sampling not
 to exceed V* in. in outside diameter,  with a
 single end opening for admission of sample.
  3.1.2  Performance Criteria.
  (a) The instrument response factors for
 each of the VOC to be measured shall be
 less  than 10. When no instrument is avail-
 able that meets this specification when cali-
 brated with the reference VOC specified  in
 the  applicable  regulation, the available in-
 strument may be calibrated  with one of the
 VOC to be measured,  or any other VOC,  so
               Pt. 60, App. A, Meth. 21

long as the instrument then has a response
factor of less than 10 for each of the VOC
to be measured.
  (b) The instrument response time shall be
equal to or less than 30 seconds. The instru-
ment pump, dilution probe (if any), sample
probe, and probe filter,  that will be used
during testing, shall all be in place during
the response time determination.
  c. The calibration precision must be equal
to or less than 10 percent of the calibration
gas value.
  d. The evaluation procedure for each pa-
rameter is given in Section 4.4.
  3.1.3 Performance  Evaluation  Require-
ments.
  a. A response factor must be determined
for each compound that  is to be measured,
either by testing or from reference sources.
The  response  factor  tests  are  required
before placing the analyzer into service, but
do not have to be repeated at subsequent in-
tervals.
  b. The calibration precision test must be
completed prior to placing the analyzer into
service, and at subsequent 3-month intervals
or at the next use whichever is later.
  c. The response time test is required prior
to placing the instrument into service. If a
modification to the sample pumping system
or flow configuration is made  that would
change the response time, a new  test is re-
quired prior to further use.
  3.2  Calibration Gases. The monitoring in-
strument is calibrated in terms of parts per
million by volume  (ppmv) of the  reference
compound specified in the applicable  regu-
lation. The  calibration  gases required for
monitoring  and  instrument  performance
evaluation are a zero gas (air, less than 10
ppmv VOC) and a calibration gas in air mix-
ture approximately equal to the leak defini-
tion specified  in the regulation. If cylinder
calibration gas mixtures are used, they must
be analyzed and certified by the  manufac-
turer to be within ±2 percent accuracy, and
a  shelf life  must  be specified.  Cylinder
standards  must be either reanalyzed or re-
placed at the end of the specified  shelf life.
Alternately, calibration  gases may be pre-
pared by the user according to any accepted
gaseous standards  preparation  procedure
that will yield a mixture accurate to within
±2 percent. Prepared standards must be re-
placed each day of use unless it can be dem-
onstrated  that degradation does  not  occur
during storage.
  Calibrations may be performed using a
compound other than the reference com-
pound if a conversion factor is determined
for that alternative compound so that the
resulting meter readings during source sur-
veys can  be  converted to  reference com-
pound results.
4. Procedures"
                                       1021

-------
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
Pt. 60, App. A, Meth. 21

  4.1  Pretest Preparations. Perform the in-
strument evaluation  procedures given  in
Section 4.4  if the evaluation  requirements
of Section 3.1.3 have not been met.
  4.2  Calibration Procedures. Assemble and
start up the VOC analyzer according to the
manufacturer's instructions. After  the  ap-
propriate warmup period and  zero internal
calibration procedure, introduce the calibra-
tion gas into the instrument sample probe.
Adjust the instrument meter readout to cor-
respond to the calibration gas value.
  NOTE: If the meter readout cannot be  ad-
justed to the proper value, a malfunction of
the analyzer is indicated and corrective ac-
tions are necessary before use.
  4.3  Individual Source Surveys.
  4.3.1   Type I—Leak Definition  Based on
Concentration. Place the probe inlet at the
surface  of the component interface where
leakage could occur. Move the probe along
the interface periphery  while observing the
instrument readout. If  an  increased meter
reading is observed, slowly sample the inter^
face where  leakage is  indicated  until the
maximum meter reading is obtained. Leave
the probe inlet at this maximum reading lo-
cation for approximately two  times the in-
strument response  time. If the maximum
observed meter reading  is greater than the
leak definition in the applicable regulation,
record and report the results as specified in
the regulation reporting requirements. Ex-
amples  of the application of this  general
technique to specific equipment types are:
  a. Valves—The most  common source  of
leaks from valves is at the seal between the
stem and housing.  Place the  probe at the
interface where the stem exits the  packing
gland and sample the stem circumference.
Also, place the probe at  the interface of the
packing  gland  take-up flange  seat  and
sample  the  periphery.  In  addition, survey
valve housings of multipart assembly at the
surface  of all interfaces where a leak could
occur.
  b. Flanges and  Other Connections—For
welded flanges, place the probe at the outer
edge of the  flange-gasket interface  and
sample  the   circumference  of the flange.
Sample other types of nonpermanent joints
(such as threaded connections) with a simi-
lar traverse.
  c. Pumps and Compressors—Conduct a cir-
cumferential traverse at the outer surface
of the  pump or compressor shaft and seal
interface. If  the source is a rotating shaft.
position the probe inlet within 1 cm of the
shaft-seal interface for the survey. If the
housing configuration prevents a complete
traverse of the shaft periphery, sample  all
accessible portions. Sample all other joints
on the pump or compressor housing where
leakage could occur.
  d. Pressure Relief Devices—The configura-
tion of most pressure relief devices prevents
          40 CFR Ch. I (7.1-92 Edition)

sampling at the sealing seat interface. For
those devices equipped with an enclosed ex-
tension, or horn, place the probe inlet at ap-
proximately the center of the exhaust area
to the atmosphere.
  e. Process Drains—For open drains,  place
the probe inlet at approximately the center
of the area open  to  the atmosphere. For
covered drains, place the probe at the sur-
face of the cover interface and conduct a pe-
ripheral traverse.
  f. Open-Ended Lines or Valves—Place the
probe inlet at approximately the center of
the opening to the atmosphere.
  g. Seal System Degassing Vents and Accu-
mulator Vents—Place the probe inlet at ap-
proximately the center  of the opening to
the atmosphere.
  h.  Access Door Seals—Place  the probe
inlet at the surface of the door seal inter-
face and conduct a peripheral traverse.
  4.3.2 Type II—"No Detectable Emission".
  Determine the local ambient concentra-
tion around the source by moving the probe
inlet randomly upwind and downwind at a
distance  of  one to two meters from the
source. If an interference exists with this
determination due to a nearby emission or
leak, the local  ambient concentration may
be determined at distances closer to the
source, but in no case shall the distance be
less than 25 centimeters.  Then move the
probe inlet to the surface of the source and
determine the  concentration described in
4.3.1. The difference between these concen-
trations determines whether there are no
detectable emissions. Record and report the
results as specified by the regulation.
  For those cases  where the regulation re-
quires a specific device installation, or that
specified vents  be ducted or piped to a con-
trol device, the existence of these conditions
shall be visually confirmed. When the regu-
lation also requires that no detectable emis-
sions exist, visual observations and sampling
surveys are required. Examples of this tech-
nique are:
  (a) Pump or Compressor Seals—If applica-
ble, determine  the type  of shaft seal. Pre-
form  a survey of the  local area ambient
VOC concentration and determine if detect-
able emissions exist as described above.
  (b) Seal System Degassing Vents, Accumu-
lator Vessel Vents, Pressure Relief Devices—
If applicable, observe whether or not the ap-
plicable ducting or piping exists. Also, deter-
mine  if any sources exist in the ducting or
piping where emissions could occur prior to
the control device. If the  required ducting
or piping exists and  there are  no sources
where the emissions could be vented to the
atmosphere prior to the control device, then
it is presumed  that no detectable emissions
are present. If there are sources in the duct-
ing or  piping where emissions could be
vented or sources where leaks could occur,
                                       1022

-------
  I
  I
  I
  I
  I
  I
  I
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
I
I
 Environmental Protection Agency

 the sampling surveys described in this para-
 graph shall be used to determine if detecta-
 ble emissions exist.
  4.3.3  Alternative Screening Procedure. A
 screening procedure based on the formation
 of bubbles in a soap solution that is sprayed
 on a potential leak source may be used for
 those sources that do not have continuously
 moving parts,  that do not have surface tem-
 peratures greater than the boiling point or
 less than the freezing point of the soap solu-
 tion, that do not have open areas to the at-
 mosphere  that the  soap  solution  cannot
 bridge, or  that do not exhibit evidence of
 liquid leakage. Sources that have these con-
 ditions present must  be surveyed using the
 instrument techniques of 4.3.1 or 4.3.2.
  Spray a  soap solution over all potential
 leak sources. The soap solution may  be a
 commercially  available leak detection  solu-
 tion or may be prepared using concentrated
 detergent and water.  A pressure sprayer or
 a squeeze bottle may be used to dispense the
 solution. Observe the potential leak sites-to
 determine  if any bubbles are formed. If no
 bubbles are observed, the source is pre-
 sumed to have no detectable emissions  or
 leaks  as applicable. If any bubbles are ob-
 served, the instrument techniques of 4.3.1 or
 4.3.2 shall  be  used to determine if  a leak
 exists, or if the source has detectable emis-
 sions, as applicable.
  4.4 Instrument Evaluation Procedures. At
 the beginning  of the instrument perform-
 ance evaluation test, assemble and start up
 the instrument according to the manufac-
 turer's   instructions   for   recommended
 warmup period  and  preliminary  adjust-
 ments.
  4.4.1  Response  Factor.  Calibrate  the in-
strument with the reference compound as
specified in the applicable regulation. For
 each organic species that is to be measured
during individual source  surveys, obtain or
prepare a known standard in air at a con-
centration  of approximately 80 percent  of
 the applicable leak definition unless limited
by volatility or explosivity. In these cases,
prepare a  standard  at 90 percent of the
saturation  concentration, or 70 percent  of
the lower  explosive limit,  respectively. In-
troduce this mixture to  the analyzer and
record the observed meter reading. Intro-
duce zero air  until a stable  reading is ob-
tained. Make a total of three measurements
by alternating between the known mixture
and zero air. Calculate the response factor
for each  repetition  and the  average re-
sponse factor.
  Alternatively,  if  response factors have
been published for the compounds of inter-
est for the instrument or detector type, the
response factor  determination  is not re-
quired,  and existing results may be refer-
enced.  Examples of published response fac-
tors for flame  ionization and catalytic oxi-
               Pt, 60, App. A, Meth. 22

dation detectors are included in Bibliogra-
phy.
  4.4.2 Calibration Precision. Make a total
of three measurements by alternately using
zero gas and the  specified calibration gas.
Record the meter readings.  Calculate the
average algebraic  difference  between the
meter readings and the known value. Divide
this average difference by the known cali-
bration value and mutiply by 100 to express
the resulting calibration  precision as a per-
centage.
  4.4.3 Response Time. Introduce zero gas
into the instrument  sample probe. When
the  meter  reading  has  stabilized,  switch
quickly to  the specified  calibration gas.
Measure the time from switching to when
90 percent of the final stable reading is at-
tained.  Perform this  test sequence  three
times and record the results. Calculate the
average response time.
5. Bibliography
  1.  DuBose.  D.A.,  and G.E.  Harris. Re-
sponse Factors of VOG -Analyzers at a Meter
Reading of 10,000 ppmv for Selected Organ-
ic Compounds.  U.S. Environmental Protec-
tion Agency, Research Triangle Park, NC.
Publication No. EPA 600/2-81-051. Septem-
ber 1981.
  2.  Brown. G.E., et al. Response Factors
of VOC Analyzers  Calibrated with Methane
for Selected Organic Compounds. U.S. Envi-
ronmental Protection Agency, Research Tri-
angle Park, NC. Publication No. EPA 600/2-
81-022. May 1981.
  3.  DuBose, D.A., et al. Response of Porta-
ble VOC Analyzers to Chemical Mixtures.
U.S. Environmental Protection Agency, Re-
search Triangle Park.  NC. Publication No.
EPA 600/2-81-110. September 1981.
METHOD 22—VISUAL DETERMINATION or FU-
   GITIVE   EMISSIONS   FROM   MATERIAL
   SOURCES  AND SMOKE  EMISSIONS  FROM
   FLARES

1. Introduction
  This method involves the visual determi-
nation of fugitive emissions, i.e., emissions
not emitted directly from a process stack or
duct.  Fugitive  emissions include emissions
that (1)  escape capture by process equip-
ment  exhaust hoods; (2) are emitted during
material   transfer;  (3)  are emitted  from
buildings  housing material  processing  or
handling equipment; and (4) are emitted di-
rectly from process equipment. This method
is used also to determine visible smoke emis-
sions  from  flares used for combustion  of
waste process materials.
  This method determines the  amount  of
time that any visible emissions occur during
the observation period, i.e.. the accumulated
emission time. This method does not require
that the opacity of emissions be determined.
                                       1023

-------