USEPA CONTRACT LABORATORY PROGRAM
   (V
   STATEMENT OF WORK

          FOR

   ORGANICS ANALYSIS
                              Multi-Media,  High-Concentration
                                      SOW No.  Rev.  9/88
                                     including Rev. 4/89
I    I
    or>
    a
HEADQUARTERS LIBRARY
ENVIRONMENTAL PROTECTION AGENCY
WASHINGTON, D.C. 20460

-------

-------
                              STATEMENT OF WORK


                              TABLE OF CONTENTS


EXHIBIT A:   SUMMARY OF REQUIREMENTS

EXHIBIT B:   REPORTING AND DELIVERABLES REQUIREMENTS

EXHIBIT C:   TARGET COMPOUND LIST  (TCL) AND CONTRACT REQUIRED QUANTITATION
             LIMITS (CRQL)

EXHIBIT D:   ANALYTICAL METHODS

EXHIBIT E:   QUALITY ASSURANCE/QUALITY CONTROL REQUIREMENTS

EXHIBIT F:   CHAIN-OF-CUSTODY, DOCUMENT CONTROL AND STANDARD OPERATING
             PROCEDURES

EXHIBIT G:   GLOSSARY OF TERMS
                                                                    Rev. 9/88

-------

-------
       EXHIBIT A
SUMMARY OF REQUIREMENTS
                                         Rev. 9/88

-------

-------
                                   SECTION I
                             GENERAL REQUIREMENTS
      The Contractor shall use proven instruments and techniques to identify
and measure the concentrations of volatile and extractable compounds listed
on the Target Compound List  (TCL) in Exhibit C.  The Contractor shall employ
state-of-the-art GC/MS and GC procedures to perform all analyses, including
all necessary preparations for analysis.

      In Exhibit D, the EPA provides the Contractor with the specific
analytical procedures to be used and defines the specific application of
these procedures to this contract.  This includes instructions for sample
preparation, gas chromatographic screening, mass spectrometric identification
and data evaluation.  Specific ions used for searching the mass spectral data
for each compound are included.

      The Contractor shall separate multiphase samples into single phase
units and prepare extracts and dilutions of samples.  The Contractor shall
screen extracts at an initial extract concentration.  Then, based on the
screening response, the Contractor shall use the specific analytical methods
described in Exhibit D to extract and concentrate samples to achieve the
Contract Required Quantitation Limits (CRQL) listed in Exhibit C.  Exhibit D
lists the analytical methods and starting points to be achieved for each of
the TCL compounds.                                              •       '

      During preparation, the Contractor shall fortify all single phase
units, blanks, and control matrix spikes with the surrogate spiking compounds
listed in Exhibit E.  Aliquots for volatile organics analysis shall be spiked
with the internal standard compounds listed in Exhibit E before purging.

      Additionally, for each single phase unit analyzed by GC/MS, the
Contractor shall conduct mass spectral library searches to determine the
possible identity of up to ten (10) nonsurrogate volatile components and up
to twenty (20) nonsurrogate extractable components that are not on the Target
Compound List (Exhibit C).

      Exhibit F contains chain-of-custody and sample documentation
requirements which the Contractor must follow in processing samples under
this contract, and specifies requirements for written laboratory standard
operating procedures.

      Sample analysis data, sample documentation and other deliverables shall
be reported as specified in Exhibit B.

      To ensure proper understanding of language utilized in this contract,
Exhibit G contains a glossary of terms.  When a term is used in the text
without explanation, the glossary meaning shall be applicable.
                                   A-2
Rev. 9/88

-------
      The samples to be analyzed by the Contractor are from known or
suspected hazardous waste sites and, potentially, may contain hazardous
organic and/or inorganic materials at high concentration levels.   The
Contractor should be aware of the potential hazards associated with the
handling and analyses of these samples.  It is the Contractor's
responsibility to take all necessary measures to ensure the health and safety
of its employees.
                                   A-3
Rev. 9/88

-------
                                  SECTION II
                             SPECIFIC REQUIREMENTS
A.   For each sample, the Contractor shall perform the following tasks:

     Task I:Receive and Prepare Hazardous Waste Samples.

     1.   Receive and handle samples under the  chain-of-custody procedures
          described in Exhibit F.   Documentation,  as  described therein,  shall
          be  required to show that all procedures  are being strictly followed.

     2.   Prepare samples as described in Exhibit  D.   Samples  are separated
          into single phase units  (if required) and screened to determine the
          proper dilution for GC/MS analysis.

          Extracts must be analyzed within 40 days of VTSR.

     Task II:   Analysis for Identity of Specific  Organic Compounds.

     1.   Extracts and aliquots prepared in Task I shall be analyzed by  GO  and
          GC/MS techniques given in Exhibit D for  the target compounds listed
          in  Exhibit C.

     2.   The target compounds listed in Exhibit C shall be identified as
          described in the methodologies given  in  Exhibit D.   Automated
          computer programs may be used to facilitate the identification.

  .   Task III:  Qualitative Verification of the Compounds Identified in
                Task II.

     1.   The compounds  analyzed by GC/MS techniques  and initially identified
          in  Task II shall be verified by an analyst  competent in the
          interpretation of mass spectra by comparison of the  suspect mass
          spectrum to the mass spectrum of a standard of the suspected
          compound.   Two criteria  must be satisfied to verify  the
          identifications:

          a.    Elution of the  sample component  at  the  same GC  relative
               retention  time  as the standard component,  and

          b.    Correspondence  of the sample  component  and standard component
               mass  spectra.   This  procedure requires  the use  of  multiple
               internal standards.

     2.   For establishing correspondence of the GC relative retention time
          (RRT),  the sample component  RRT must  compare within  ±0.06 RRT units
          of  the  RRT of  the standard component.  For  reference,  the
          calibration standard must be run on the  same 12-hour time period as
          the sample.
                                   A-4
Rev. 9/88

-------
For comparison of standard and sample component mass spectra,  mass
spectra obtained on the Contractor's GC/MS are required.  Once
obtained, these standard spectra may be used for identification
purposes only if the Contractor's GC/MS meets the DFTPP or BFB daily
tuning requirements of Tables 1.1 and 1.2 in Exhibit E.  The
standard spectra used may be from a laboratory generated library or
obtained from the calibration standard run used to obtain reference
RRTs.  The requirements for qualitative verification by comparison
of mass spectra are as follows:

a.   All ions present in the standard mass spectrum at a relative
     intensity greater than 10 percent (most abundant ion in the
     spectrum equals 100 percent) must be present in the sample
     spectrum.

b.   The relative intensities of ions specified in (1) must agree
     within plus or minus 20 percent between the standard and sample
     spectra.

c.   Ions greater than 10 percent in the sample spectrum but not
     present in the standard' spectrum must be considered and
     accounted for by the analyst making the comparison.  When GC/MS
     computer data processing programs are used to obtain the sample
     component spectrum, both the processed and the raw spectra must
     be evaluated. In Task III, the verification process should
     favor false positives.

If a compound analyzed.by GC/MS  techniques and initially identified
in Task II cannot be verified by all of the criteria  in items 1 and
2 above, but  in the technical judgement of the mass spectral
interpretation specialist the identification is correct, then the
Contractor shall report that identification, and proceed with
quantification in Task IV.

The Toxaphene and the Aroclor compounds listed in Exhibit C and
analyzed by GC/EC techniques shall have their identifications
verified by an analyst competent in the interpretation of gas
chromatograms.  Two criteria must be satisfied to verify the
identifications:

a.   Elution of the sample component within the retention time
     window (established by the procedures in Exhibit E) of the
     standard component analyzed on the same GC column and
     instrument,  as part of the same analytical sequence specified
     in Exhibit D ARO.

b.   Analysis of the sample and standard on a second GC column with
     a stationary phase with retention characteristics dissimilar to
     that used in a. above, and meeting the same criteria for
     elution of the sample component and the standard as in a.
     above.
                          A-5
Rev. 9/88

-------
TaskIV:  Quantification of Compounds Verified in Task III.

1.   The Contractor shall quantify components analyzed by GC/MS
     techniques and identified in Task II and verified in Task III by the
     internal standard method stipulated in Exhibit D.   Where multiple
     internal standards are required by EPA,  the Contractor shall perform
     quantitation utilizing the internal standards specified in Exhibit
     E,  Part 2, Tables 2.1 or 2.2.

2.   The Contractor shall determine response factors for each 12-hour
     time period of GC/MS analysis and shall include a calibration check
     of the initial calibration as described in Exhibit E.

3.   The Contractor shall quantify components analyzed by GC/EC
     techniques and identified in Task II and verified in Task III by the
     external standard method stipulated in 'Exhibit D ARO.

4.   The Contractor shall perform an initial three-point calibration,
     verify its linearity,  determine the degradation of labile
     components, and determine calibration factors for all standards
     analyzed by GC/EC techniques as part of an analytical sequence, as
     described in Exhibit D ARO and Exhibit E.

Task V:   Tentative Identification of Non-TCL Sample Components.

1.   For each sample,  the Contractor shall conduct mass spectral library
     searches to determine tentative compound identifications as follows.
     For each volatile fraction, the Contractor shall conduct a search to
     determine the possible'identity of the ten (10) nonsurrogate organic-
     compounds of greatest concentration which are not listed in Exhibit
     C.   For each extractable fraction, the Contractor shall conduct a
     search to determine the possible identification of the (20)
     nonsurrogate organic compounds of greatest concentration which are
     not listed in Exhibit C.  In performing searches,  the 1985 (or most
     recent) release of the National Bureau of Standards library
     (containing 42,261 spectra) must be used.  NOTE:  Substances with
     responses less than 10 percent of the nearest internal standard are
     not required to be searched in this fashion.

     Only after visual comparison of sample spectra with the spectra from
     the library searches will the mass spectral interpretation
     specialist assign a tentative identification.  If the compound does
     not meet the identification criteria of Task III,  it shall be
     reported as unknown.  The mass spectral specialist should give
     additional classification of the unknown compound, if possible
     (i.e.,  unknown aromatic,  unknown hydrocarbon, unknown acid type,
     unknown chlorinated compound).   If probable molecular weights can be
     distinguished, include them.

     The Contractor shall not report as tentatively identified compounds
     (TIC)  any TCL compounds  from another analytical fraction (i.e.,  do
     not report late eluting  volatile compounds as TICs in the
     semivolatile analysis).
                              A-6                              Rev.  9/88

-------
Task VI:  Quality Assurance/Quality Control Procedures.

1.   All specific quality assurance procedures prescribed in Exhibit E
     shall be strictly adhered to by the Contractor.   Records documenting
     the use of the protocol shall be maintained in accordance with the
     document control procedures prescribed in Exhibit F, and shall be
     reported in accordance with Exhibit B, Reporting Requirements and
     Deliverables.   Single phase units,  method blanks and control matrix
     spikes shall be carried through the entire analytical process from
     extraction, GC screen, to final GC/MS analysis,  including all data
     reporting requirements and magnetic tape data storage.

2.   The Contractor shall perform one control matrix spike sample
     analysis for each case received, or for each 20 single phase units,
     or each 14 calendar day period during which single phase units in a
     Case were received (said period beginning with the receipt of the
     first sample in that Sample Delivery Group) whichever is most
     frequent.

     Samples, blanks and control matrix spikes shall be carried through
     the entire analytical process from extraction to final GC/MS
     analysis, including all Contract Performance/Delivery Requirements
     (see Contract Schedule).

3.   The Contractor shall prepare and analyze one laboratory reagent
     blank (method blank) for each group once for:

     o each Case, of single phase units received, OR

     o each 20  single phase units  in a Case, OR

     o each 14  calendar day period during which single phase units in a
       Case were received  (said period beginning with the  receipt of the
       first  sample in that Sample Delivery Group), OR

     o whenever samples are extracted,

       whichever is most frequent

     Volatile analysis requires one method blank for each 12-hour time
     period when volatile TCL compounds are analyzed.

     Extractable method blanks shall be carried through the entire
     analytical process from extraction to final GC/MS or GC/EC analysis,
     including all Contract Performance/Delivery Requirements (see
     Contract Schedule).

4.   The Contractor shall perform instrument calibration (by "hardware
     tune") for each 12-hour time period, to include:
     decafluorotriphenylphosphine (DFTPP) and/or bromofluorobenzene (BFB)
     as applicable, and a specific calibration using standards of defined
     concentration to monitor response,  retention time and mass spectra.

     Additional quality control shall be conducted in the form of the
     analysis of Performance Evaluation check samples submitted to the
     laboratory by EPA.   The results of comparison studies are due within
                               A-7
Rev. 9/88

-------
     40 calendar days of receipt of the samples.  The results of all such
     control or PE check samples may be used as grounds for termination
     of noncompliant contractors.

EPA has provided to the Contractor formats for the reporting of data
(Exhibit B).  The Contractor shall be responsible for completing and
returning analysis data sheets in the format specified in this SOW and
within the  time specified in the Contract Performance/Delivery Schedule.

1.   Use of formats other than those designated by EPA will be deemed as
     noncompliance.   Such data are unacceptable.  Resubmission in the
     specified format at no additional cost to the government will be
     required.

2.   Computer generated forms may be submitted in the hardcopy data
     package(s) provided that the  forms are in EXACT EPA FORMAT.   This
     means that the order of data  elements is the same as on each EPA
     required form,  including form numbers and titles, page numbers and
     header information

The Contractor shall provide analytical equipment and technical
expertise for this contract as specified following:

1.   The Contractor shall have sufficient gas chromatograph (CC)  and gas
     chromatograph/mass spectrometer/data system (GC/MS/DS) capability to
     meet all the terms and conditions of the Contract.   Instrument
     requirements are defined in Section III, Detailed Technical  &
     Management Requirements.  The  Contractor shall maintain,  at a
     minimum,  all analytical equipment allocated for this contract at the
     time of contract award.

2.   The Contractor's instrument systems shall have the following:

     a.    The  GC/MS  shall be equipped with a glass jet separator  when
          using packed columns.

     b.    The  computer shall be  interfaced by hardware to the mass
          spectrometer and be capable of acquiring continuous mass  scans
          for  the duration of the  chromatographic program.

     c.    The  computer shall, be  equipped with mass storage devices  for
          saving all  data from the GC/MS runs.

     d.    Computer software shall  be available to allow searching GC/MS
          runs for specific ions and plotting the intensity of the  ions
          with respect to time or  scan number.

     e.    The  GC/MS  shall be equipped with a split/splitless  injector and
          GC to MS  interface capable of extending a fused silica
          capillary  column into  the  ion source.   The column is to be  30
          meters long by 0.25  or 0.32 mm inside  diameter,  bonded  DB-5,
          fused silica or equivalent.
                              A-8
Rev. 9/88

-------
     f.   The GG  for Aroclor analysis  shall be equipped with dual wide
         bore  (>0.53 nun) capillary columns and  a  suitable  detector  as
         described in  Exhibit  D.  The instrument  must be capable of
         operating with a  temperature program.

3.   The Contractor shall use  a magnetic tape storage device capable of
     recording  data and suitable for  long-term,  off-line  storage.   The
     Contractor shall  retain all raw  GC/MS data  acquired  under this
     contract on magnetic  tape  in appropriate instrument  manufacturer's
     format.  The Contractor is required to retain the magnetic tapes
     with associated hardcopy  tape logbook identifying tape contents (see
     Exhibit  B) for 365 days after data submission.   During that time,
     the Contractor shall  submit tapes and logbook within 7 days of
     request, as specified in  the Contract Performance/Delivery Schedule.

4.   The Contractor shall have a computerized MS library  search system
     capable  of providing  a forward comparison,  utilizing the  standard
     spectra  contained in  the  mass spectral library.   The 1985 (or most
     recent)  release  of the National  Bureau of Standards  library
     (containing 42,261 spectra) must be used.

     a.   The system shall  provide a  numerical ranking  of the  standard
         spectra most  closely  corresponding  to  the sample  spectra
         examined.

     b.   The data system shall have  software capable of  removing
         background signals from spectra.

5.   The Contractor shall  have, in-house and operable,  a  device capable
     of analyzing purgeable organics  as described in  Exhibit D.

The minimum functional requirements necessary to meet the terms and
conditions of this contract are listed below.  The Contractor shall
designate and utilize key personnel to perform these  functions.  The EPA
reserves the right to review personnel qualifications and experience.
See Section III, Detailed Technical & Management Requirements.

o   GC/MS/DS  operation.
o   Mass spectral interpretation.
o   Sample  extraction  and  concentration.

o   Purge and trap volatile organic compounds analysis.
o   Pesticide residue  analysis  of organochlorine pesticides and PCBs,
    including clean-up procedures.

o   Quality assurance/quality  control
o   Sample  receipt, storage, and tracking,  including  chain-of-custody
    procedures.
The Contractor shall respond in a timely manner to requests from data
recipients for additional information or explanations that result from
the Government's  inspection activities.
                               A-9
. Rev. 9/88

-------
F.   The Contractor shall preserve all sample extracts after analysis in
     bottles/ vials with teflon-lined septa and shall maintain stored
     extracts at 4'C (±2*C).   The Contractor is required to retain the sample
     extracts for 365 days after data submission.   During that time,  the
     Contractor shall submit the single phase unit extracts within 7  days
     after request, as specified in the Contract Performance/Delivery
     Schedule.

G.   The Contractor shall adhere to chain-of-custody procedures described in
     Exhibit F.  Documentation,  as described therein, shall be required to
     show that all procedures are being strictly followed.   This
     documentation shall be reported as the complete Case file purge  (see
     Exhibit fl).

H.   Sample shipments to the Contractor's facility will be scheduled  and
     coordinated by the EPA CLP Sample Management Office (SMO) acting on
     behalf of the Project Officer.  The Contractor shall communicate with
     SMO personnel by telephone as necessary throughout the process of sample
     scheduling, shipment, analysis and data reporting, to ensure that
     samples are .properly processed.

     If there are problems with the samples (e.g., mixed media, containers
     broken or leaking) or sample documentation/paperwork (e.g., Traffic
     Reports not with shipment,  sample and Traffic Report numbers do  not
     correspond) the Contractor shall immediately contact SMO for resolution.
     The Contractor shall immediately notify SMO regarding any problems and
     laboratory conditions that affect the timeliness of analyses and data
     reporting.  In particular,  the Contractor shall notify SMO personnel in
     advance regarding sample data that will be delivered late and shall
     specify the estimated delivery date.

I.   Sample analyses will be  scheduled by groups of samples,  each defined as
     a Case and identified by a unique EPA Case number assigned by SMO.  A
     Case signifies a group of samples collected at one site  or geographical
     area over a finite time  period, and will include one or  more field
     samples with associated blanks.  Samples may  be shipped  to the
     Contractor in a single shipment or multiple shipments  over a period of
     time,  depending on the size of the Case.

     A Case consists of one or more Sample Delivery Group(s).   A Sample
     Delivery Group (SDG) is  defined by the following,  whichever is most
     frequent:

     o  each Case  of single phase units received, OR

     o  each 20 single phase units within a Case, OR
     o  each 14 calendar day period during which single phase units in a Case
        are received (said period beginning with the receipt of  the first
        sample  in  the Sample Delivery Group).

     Data for all samples in  a Sample Delivery Group are due  concurrently 40
     days after receipt of the last sample received in the  Sample Delivery
     Group.   Data for all samples in a Sample Delivery Group  must be
     submitted together (in one  package)  in the order specified in Exhibit B.


                                   A-10                            Rev.  9/88

-------
M.
The Sample Delivery Group number is the EPA sample number of the first
sample received in the SDG.  When several samples are received together
in the first SDG shipment, the SDG number shall be the lowest sample
number (considering both alpha and numeric designations) in the first
group of samples received under the SDG.  The SDG number is reported on
all data reporting forms.

The SDG Receipt Date is the day the last sample in the SDG is received.
Data for all samples in the SDG are due 40 days following this date.

The Contractor is responsible for identifying each Sample Delivery Group
as samples are received, through proper sample documentation (see
Exhibit B) and communication with SMO personnel.

Each sample received by the Contractor will be labeled with an EPA
sample number, and accompanied by a Traffic Report form bearing the
sample number and descriptive information regarding the sample.  The
Contractor shall complete and sign the Traffic Report, recording the
date of sample receipt and sample condition on receipt for each sample
container.

The Contractor shall submit signed copies of Traffic Reports for all
samples in a Sample Delivery Group to SMO within 3 calendar days
following receipt of the last sample in the Sample Delivery Group.
Traffic Reports shall be submitted in Sample Delivery Group sets (i.e.,
all Traffic Reports for a Sample Delivery Group shall be clipped
together) with an SDG Cover Sheet containing information regarding the
Sample Delivery Group, as specified in Exhibit B.

EPA Case numbers (including SDG numbers) and EPA sample numbers shall be
used by the Contractor in identifying samples received under this
contract both verbally and in reports/correspondence.

Samples will routinely be shipped to the Contractor through an overnight
delivery service.  However, as necessary, the Contractor shall be
responsible for any handling or processing required for the receipt of
sample shipments, including pick-up of samples at the nearest servicing
airport, bus station or other carrier service within the Contractor's
geographical area.  The Contractor shall be available to receive sample
shipments at any time the delivery service is operating, including
Saturdays.

The Contractor shall accept all samples scheduled by SMO, provided that
the total number of samples received in any calendar month does not
exceed the monthly limitation expressed in the contract.  Should the
Contractor elect to accept additional samples, the Contractor shall
remain bound by all contract requirements for analysis of those samples
accepted.
                                   A-ll
                                                               Rev. 9/88

-------
                                 SECTION  III


                 DETAILED TECHNICAL & MANAGEMENT REQUIREMENTS


As cited in Section II, Task VI,  the Contractor shall have the following
technical and management capabilities:

A.   TECHNICAL CAPABILITY

     1.   Technical Functions

          a.   GC/MS Laboratory Supervisor

               (1)  Responsible for all technical efforts of  the GC/MS
                   laboratory to meet all terms and conditions of the EPA
                   contract.

               (2)  Qualifications:

                   (a)  Education:

                        Minimum of Bachelor's degree in chemistry or any
                        physical science.

                   (b)  Experience:

                        Minimum.of three years of laboratory experience,
                        including at least one year of supervisory
                        experience.

          b.   GC/MS Operator Qualifications

               (1)  Education:

                   Minimum of Bachelor's degree in chemistry or any physical
                   science.

               (2)  Experience:

                   One -year of experience in operating and maintaining
                   GC/MS/DS with degree in chemistry or a physical science,
                   or three years of experience in operating and maintaining
                   GC/MS/DS.

          c.   Mass Spectral Interpretation Specialist Qualifications

              (1)  Education:

                   o   Minimum of Bachelor's  degree  in chemistry  or  any
                       physical science.

                   o   Training course(s)  in  mass  spectral  interpretation.


                                   A-12                             Rev. 9/88

-------
     (2)   Experience:

         . Minimum of two  years  of  experience.

d.   GC Laboratory Supervisor

     (1)   Responsible for all technical  efforts  of the  GC
          laboratory.

     (2)   Qualifications:

          (a)  Education:

              Minimum of  Bachelor's degree  in chemistry or  any
              physical science.

          (b)  Experience:

              Minimum of  three years of laboratory experience,
              including at least  one year of supervisory
              experience.

e.   Pesticide Residue Analysis Expert Qualifications

     (1)   Education:

          Minimum of Bachelor's  degree in chemistry or  any physical
          science.

     (2)   Experience:

          Minimum of two  years  of  experience in  operating and
          maintaining GC  and  interpreting GC-chromatograms.

f.   Sample Preparation  Laboratory Supervisor

     (1)   Responsible for  all technical  efforts  of sample
          preparations to  meet  all terms and conditions of the EPA
          contract.

     (2)   Qualifications:

          (a)  Education:

              Minimum of  Bachelor's degree  in chemistry or  any
              physical science.

          (b)  Experience:

              Minimum of  three years of laboratory experience,
              including at least  one year of supervisory
              experience.
                         A-13                             Rev. 9/88

-------
     h.
Extraction/Concentration Expert Qualifications

(1)  Education:

     Minimum of High school diploma and knowledge of general
     chemistry.

(2)  Experience:

     Minimum of one year of experience.

Technical Staff Redundancy

The bidder shall have a minimum of one (1) chemist available at
any one time as a back-up technical person with the following
qualifications, to ensure continuous operations to accomplish
the required work as specified by EPA contract.

(1)  Education:

     Minimum of Bachelor's degree in chemistry or any physical
     science.

(2)  Experience: Minimum-of one year in each of the following
     areas -

     o  GC/MS operation and maintenance for volatiles and
        semivolatiles analyses.
              o
              o
              o
        Mass spectral interpretation.
        Extraction.
        Pesticide analysis.
2.   Facilities
     The adequacy of the facilities and equipment is of equal importance
     as the technical staff to accomplish the required work as specified
     by the EPA contract.

     a.   Sample  Receipt Area

          Adequate,  contamination-free,  well  ventilated work space
          provided with  chemical  resistant bench  top for receipt  and safe
          handling of EPA  samples.

     b.   Storage Area

          Sufficient refrigerator space to maintain unused EPA sample
          volume  for 60  days after  data submission and sample extracts
          for 365 days after data submission.   NOTE:  Volatiles.
          semivolatiles. extracts,  and standards  must each be stored
          separately.
                              A-14
                                                     Rev. 9/88

-------
3.
c.   Sample Preparation Area

     Adequate,  contamination-free,  well-ventilated work space
     provided with:

     (1)   Benches  with  chemical  resistant  tops,  exhaust hoods.
          Note:  Standards must be prepared in a  glove  box or
          isolated area.

     (2)   Source of  distilled or demineralized organic-free water.

     (3)   Analytical balance(s)  located away from draft and rapid
          change in  temperature.

Instrumentation

At a minimum, the Contractor shall have the following instruments
operative at the time of the Preaward Site Evaluation and committed
for the full duration of the contract.

a.   Primary Instrument Requirements

     (1)   60 Phase Units/Month Capacity
Purpose
Analysis


GPC Cleanup
Screening
I
Fraction
Volatiles
Extractables
Aroclors/
Toxaphene
Extractables

No. of
Instrument(s)
1
1
1
1
1
| Type of
| Instrument
1
1
JGC/MS/DS with
j purge and trap
| device
1
IGC/MS/DS
l
IGC/EC with
jdual column
i
JGPC with UV
| detector
1
IGC/FID
1
         Note:   For  contracts of two  (2) bid lots or more:

         o   Minimum of three (3)  GC/MS/DS and three (3) GC systems are
             required at the time of on-site laboratory evaluation.

         o   An additional one (1)  GC/MS/DS and one (1) GC system with
             dual detectors are required as a back-up system at the time
             of on-site laboratory evaluation.
                              A-15
                                                          Rev. 9/88

-------
     4.
          b.  Secondary Instrument Requirements

              (1)  60 Phase Units/Month Capacity

                   The.Contractor shall have the following instruments in
                   place and operational at any one time as a back-up system;
                   Quantity

                   One
                   One
                   One
                                Instruments

                                GC/MS/DS
                                Purge and Trap Device
                                GC with dual detectors (FID and EC)
                   These instruments must be included in the bidder's
                   inventory of equipment along with those in (1) above.

                   In addition, the Contractor shall have an in-house stock
                   of instrument parts and circuit boards to ensure
                   continuous operation to meet contract-specified holding
                   and turnaround times.

               Instrument Specifications

               Instrument specifications are described in detail  in the
               Statement of Work  (SOW)  in  the following Exhibits.
     o Purge and trap device

     o GC/MS/DS

     o GC
Data Handlingand Packaging
                                          Exhibit D
                                          Exhibits A and D

                                          Exhibit D
          The Contractor shall  be  able  to  submit  reports  and data packages  as
          specified in the  Statement  of Work Exhibit B.   To  complete this
          task,  the Contractor  shall  be required  to:

          a.   Provide  space, tables  and copy machines to meet  the contract
              requirements.

          b.   Designate  personnel.

B.   LABORATORY MANAGEMENT  CAPABILITY

     The Contractor must have an organization with well-defined
     responsibilities for each individual  in the  management  system to ensure
     sufficient resources for EPA contract(s)and  to maintain a  successful
     operation.   To establish this capability,  the Contractor shall  designate
     personnel to carry out the following  responsibilities for  the EPA
     contract.  Functions include, but  are not limited to, the  following:
                                   A-16
                                                          Rev. 9/88

-------
Technical Staff

Responsible for all technical efforts for the EPA contract.

Project Manaeer

Responsible for overall aspects of EPA contract(s) (from sample
receipt through data delivery) and shall be the primary contact for
EPA Headquarters Project Officer and Regional Deputy Project
Officers.

Sample Custodian

Responsible for receiving the EPA samples (logging, handling and
storage).

Quality Assurance Officer

Responsible for overseeing the quality assurance aspects of the data
and reporting directly to upper management.

Data Reporting and Delivery Officer

Responsible for all aspects of data deliverables:   organization,
packaging, copying, and delivery.
                          A-17
Rev. 9/88

-------
               EXHIBIT B
REPORTING AND DELIVERABLES REQUIREMENTS

-------

-------
                                  EXHIBITB
                   REPORTING AND DELIVERABLES REQUIREMENTS
SECTION I:

SECTION II:



SECTION III:

SECTION IV:
                                                Page  No.


Contract Reports/Deliverables Distribution	   B-2

Report Descriptions and Order of Data

Deliverables 	   B-5

Forms Instruction Guide 	   B-21

Data Reporting Forms 	   B-40
                                   B-l
                                                  Rev.  9/88

-------
                                  SECTION I

                  CONTRACT REPORTS/DELIVERABLES DISTRIBUTION
      The following table reiterates the Contract reporting and deliverables
requirements specified in the Contract Schedule and specifies the
distribution that is required for each deliverable.  NOTE: Specific recipient
names and addresses are subject to change during the term of the contract.
The Project Officer will notify the contractor in writing of such changes
when they occur.
         Item
    No.
  Conies
    Delivery
    Schedule
                                                   Distribution
  A. Contract Start-Up   2
     Plan

  B. Updated SOPs        1
         Item
  No.
Copies
 7 Days after
 contract award

 120 days after
 contract award

Delivery
Schedule	
                                                       Distribution
                                (5)
                  (6)
 **C. Sample Traffic   1
      Reports
***D. Sample Data      1
      Summary Package
      last sample
      in SDG

***£. Sample Data      3
      Package
  F. GC/MS Tapes
 Lot
         3 days after
         receipt of
         last sample
         in Sample
         Delivery Group
         (SDG)****

         35 days after
         receipt of
35 days after    X
receipt of
last sample
in SDG

Retain for 365
days after data
submission, or
submit within 7
days after
receipt of
written request
by PO and/or EMSL/LV.
As Directed
                                    B-2
                                           Rev. 1/89

-------
         Item
  No.
 Copies
Delivery
Schedule
                                                       Distribution
JAL
(5)
  G. Extracts
  H. Complete Case
     File Purge
Lot     Retain  for 365
        days after data
        submission, or
        submit  within 7
        days after
        receipt of
        written request
        by  PO or  SMO

  1      Submit  180 days
.Pkg     after data
        submission or 7
        days after receipt
        of  written request
        by  PO or  SMO.
                      As Directed
* Contractor must be prepared to receive samples within 30 days of contract
award.  NOTE:  EPA cannot guarantee exact adherence to start-up plan that is
agreed upon by the PO and Contractor, but will attempt to meet it as close as
possible.

** Also required in the Sample Data Package.

*** Concurrent delivery required-.  Delivery shall be made such that all
designated recipients receive the item on the same"~calendar day.
**** Sample Delivery Group (SDG) is a group of samples within a Case,
received over a period of 14 days or less and not exceeding 20 samples.
for all samples in the SDG are due concurrently.  (See SOW Exhibit A,
paragraph J., for further description).
                                                    Data
NOTE:  As specified in the Contract Schedule (G.6 Government Furnished
Supplies and Materials), unless otherwise instructed by the CLP Sample
Management Office, the Contractor shall dispose of unused sample volume and
used sample bottles/ containers no earlier than sixty (60) days following
submission of analytical data.

Distribution Addresses:

(1)  USEPA Analytical Operations Branch (WH 548A)
     401 M Street, SW
     Washington,  DC  20460
     ATTN:   (Project Officer's Name)

(2)  USEPA
     Contracts Management Division (MD-33)
     Administration Building Lobby,  Alexander Drive
     Research Triangle Park,  NC  27711
     ATTN:   (Contract Officer's Name)
                                   B-3
                                           Rev. 9/88

-------
(3)   USEPA Contract Lab  Program
     Sample Management Office  (SMO)
     P.  0.  Box 818
     Alexandria,  VA  22313

     For overnight delivery service,  use street address:
     209 Madison  Street,  Suite 200
     Alexandria,  VA  22314

(4)   USEPA Environmental Monitoring
     Systems  Laboratory  (EMSL-LV)
     P.  0.  Box 15027
     Las Vegas, NV  89114
     ATTN:  Data  Audit Staff

     For overnight delivery service,  use street address:
     944 E. Harmon, Executive  Center
     Las Vegas, NV  89109
     ATTN:  Data  Audit Staff

(5)   USEPA REGIONS:

      The CLP Sample Management Office acting on behalf of the Project
      Officer, will provide the Contractor with the list of addressees for
      the ten EPA Regions.  SMO will  provide the Contractor with updated
      Regional address/name lists  as  necessary throughout the period of the
      contract and identify other  client recipients on a .case-by-case basis.

(6)   NEIC,  Contractor Evidence Audit  Team
     12600 West Coifax,  Suite  310
     Lakewood, Colorado  80215
                                   B-4                          Rev.  9/88

-------
                                  SECTION II
              REPORT DESCRIPTIONS AND ORDER OF DATA DELIVERABLES
The Contractor laboratory shall provide reports and other deliverables as
specified in the Contract Schedule (Performance/Delivery Schedule, Section
F.I). The required content and form of each deliverable is described in this
Exhibit.

All reports and documentation MUST BE:

      o  Legible,
      o  Clearly  labeled and  completed in  accordance with  instructions  in  this
         Exhibit,
      o  Arranged in  the order  specified in  this Section,  and

      o  Paginated.

If submitted documentation does not conform to the above criteria, the
Contractor will be required to resubmit such documentation with
deficiency(ies) corrected, at no additional cost to the Agency.

Whenever the Contractor  is required to submit or resubmit data as a result of
an on-site laboratory evaluation or through a PO/DPO action, the data must be
clearly marked as ADDITIONAL DATA and must be sent to all three contractual
data recipients  (SMO, EMSL-LV, and Region).   A cover letter shall be included
which describes what data is being delivered, to which EPA Case(s) it
pertains, and who requested the data.

Whenever the Contractor  is required to submit or resubmit data as a result of
Contract Compliance Screening (CCS) review by SMO, the data must be sent to
all three contractual data recipients (SMO,  EMSL/LV and Region), and in all
three instances must be  accompanied by a color-coded COVER SHEET (Laboratory
Response To Results of Contract Compliance Screening)  provided by SMO.

Section III of this Exhibit contains copies  of the required data reporting
forms in Agency-specified formats, along with instructions to assist the
Contractor in accurately providing the Agency all required data. •

Descriptions of the requirements for each deliverable  item cited in the
Contract Performance/Delivery Schedule (Contract Schedule, Section F.I) are
specified in parts A-G of this Section.  Items submitted concurrently MUST BE
arranged in the order listed.  Additionally, the components of each item MUST
BE arranged in the order presented in this Section when the item is
submitted.

Examples of specific data deliverables not included herein may be obtained by
submitting a written request to the EPA Project Officer,  stating the
information requested,  and signed by the  Laboratory Manager.
                                   B-5
Rev. 9/88

-------
A.   Contract Start-Up Plan

     The Contractor shall submit a contract start-up plan for EPA approval as
     specified in the Contract Performance/Delivery Schedule.  The plan shall
     set forth the Contractor's proposed schedule for receiving samples
     starting with the 30th calendar day after award and ending with the date
     the Contractor is capable of receiving the full monthly sample allotment
     stipulated in the Contract.  The Project Officer will review the
     contract start-up plan within 7 days of submission and will notify the
     Contractor of the plan's status.

     NOTE:  The Contractor shall be required to receive samples within 30
     days of contract award.  EPA can't guarantee exact adherence to start-up
     plan that is agreed upon by the PO and Contractor, but will attempt to
     meet it as close as possible.

B.   Updated SOPs

     The Contractor shall submit updated copies of all required Standard
     Operating Procedures (SOPs) that were submitted with the prebid
     Performance Evaluation sample results.  The updated SOPs must address
     any and all issues of laboratory performance and operation identified
     through the review of the Performance Evaluation sample data and the
     evaluation of Bidder-Supplied Documentation.

     The Contractor must supply SOPs for :

     1.   Sample receipt and logging.

     2.   Sample and extract storage.

     3.   Preventing sample contamination.

     4.   Security for laboratory and samples.

     5.   Traceability/Equivalency of standards.

     6.   Maintaining instrument records and logbooks.

     7.   Sample analysis and data control systems.

     8,   Glassware cleaning.

     9.   Technical and managerial review of laboratory operation and data
          package preparation.

     10.  Internal review of contractually-required quality assurance and
          quality control data for each individual data package.

     11.  Sample analysis, data handling and reporting.

     12.  Chain-of-custody.

     13.  Document control, including case file preparation.

                                    B-6                         Rev. 9/88

-------
C.   Sample Traffic Reports

     Original Sample Traffic Report page marked "Lab Copy for Return to SMO"
     with lab receipt information and signed in original Contractor
     signature, for each sample in the Sample Delivery Group.

     Traffic Reports (TRs) shall be submitted, in Sample Delivery Group (SDG)
     sets (i.e., TRs for all samples in an SDG shall be clipped together),
     with an SDG Cover Sheet attached.

     The SDG Cover Sheet shall contain the following items:

     o   Lab  name
     o   Contract  number

     o   Sample  Analysis Price  -  full  sample  price  from contract.

     o   Case Number

     o   List of EPA  sample  numbers  of all  samples  in  the  SDG,  identifying the
         first  and last  samples  received, and their dates  of  receipt  (LRDs).
         NOTE: When more than one sample  is received in the first  or  last  SDG
         shipment,  the "first"  sample  received would be the lowest sample
         number  (considering both alpha and numeric designations);  the  "last"
         sample  received would be the  highest sample number (considering both
         alpha  and numeric designations).

     In addition,  each Traffic Report must be clearly marked  with the SDG'
     Number,  the sample  number of the first sample in the SDG (as described
     in the following paragraph).   This  information should be entered below
     the Lab Receipt Date on the TR.  In addition,  the TR for the last sample
     received in the SDG must be clearly marked "SDG - FINAL  SAMPLE."

     The EPA sample number of the first  sample received in the SDG is the SDG
     number.   When several samples  are received together in the first SDG
     shipment,  the SDG number shall be the lowest sample number (considering
     both alpha and numeric  designations) in the first group  of samples
     received  under the SDG.   (The SDG  number is also reported on all data
     reporting forms.  See Section  III,  Forms Instruction Guide.)

     If samples are received at the laboratory with multi-sample Traffic
     Reports, all  the samples on one multi-sample TR may not  be necessarily
     in the same SDG.  In this instance,  the  laboratory must  make the
     appropriate number  of photocopies of the TR,  and submit  one copy with
     each SDG cover sheet.

D.   Sample Data Summary Package

     As specified  in the Delivery Schedule,  one  Sample Data Summary Package
     shall be delivered  to SMO concurrently with delivery of  other required
     sample data.   The Sample  Data  Summary  Package  consists of copies of
     specified  items  from the  Sample Data Package.   These  items are listed
     below and  described under  part  C, Sample Data  Package.
                                   B-7                          Rev.  9/88

-------
     The Sample Data Summary Package shall be ordered as follows and shall be
     submitted separately (i.e.,  separated by rubber bands,  clips or other
     means)  directly preceding the Sample Data Package.   Sample data forms
     shall be arranged in increasing EPA sample number order,  considering
     both letters and numbers.  BE400 is a lower sample number than BF100, as
     E precedes F in the alphabet.

     The Sample Data Summary Package shall contain data for  samples in one
     Sample  Delivery Group of the Case,  as follows:

     1.    Case Narrative

     2.    By fraction (HCV,  HCE,  and HCA,)  and by phase  unit within each
          fraction -  tabulated target compound results  (Form I)  and
          tentatively identified  compounds  (Form I,  TIC)  (HCV,  HCE only).

     3.    By fraction (HCV,  HCE,  HCA)  -  surrogate spike  analysis results (Form
          II).

     4.    By fraction (HCV,  HCE,  HCA)  -  control matrix  spike results (Form
          III)

     5.    By fraction (HCV,  HCE,  HCA)  -  blank data (Form IV) and tabulated
          results (Form I)  including tentatively identified  compounds  (Form I,
          TIC)(HCV, HCE only).

E.   Sample  Data Package

     The Sample Data Package is divided into the five major  units described
     below.  The last three units  are each specific to an analytical fraction
     (volatiles, extractables, Aroclors).  If the analysis of  a fraction is
     not required, then that fraction-specific unit is  not required as a
     deliverable.

     The Sample Data Package shall include data for analyses of all samples
     in one  Sample Delivery Group, including field samples,  reanalyses,
     blanks, and control matrix spikes.

     1.    Case Narrative

          This document shall be  clearly labeled "Case  Narrative" and  shall
          contain:  laboratory name;  Case number;  sample numbers in the  Sample
          Delivery Group (SDG) and the phase units analyzed  from"each  sample,
          differentiating between initial analyses and  re-analyses;  SDG
          number; Contract number; and detailed documentation  of any quality
          control, sample,  shipment and/or analytical problems encountered in
          processing the samples  reported in the data package.

          Whenever data from sample re-analyses are  submitted,  the Contractor
          shall state in the Case Narrative for each re-analysis,  whether  it
          considers the re-analysis to be billable,  and  if so,  why.
                                   B-8                          Rev. 9/88

-------
     The Contractor must also include any problems encountered; both
     technical and administrative, the corrective actions taken, and
     resolution.

     The Case Narrative shall contain the following statement, verbatim:
     "I certify that this data package is in compliance with the terms
     and conditions of the contract,  both technically and for
     completeness, for other than the conditions detailed above.  Release
     of the data contained in this hardcopy data package has been
     authorized by the Laboratory Manager or his designee,  as verified by
     the following signature."  This  statement shall be directly followed
     by signature of the laboratory Manager or his designee with a typed
     line below it containing the signer's name and title,  and the date
     of signature.

     Additionally, the Case Narrative itself must be signed in original
     signature by the Laboratory Manager or his designee and dated.

2.   Traffic Reports

     A copy of the Sample Traffic Reports submitted in Item A for all of
     the samples in the SDG.   The Traffic Reports shall be  arranged in
     increasing EPA sample number order,  considering both letters and
     numbering in ordering samples.   Copies of the SDG cover sheet are to
     be included with the copies of the Traffic Reports.

     If samples are received at the laboratory with multi-sample Traffic
     Reports (TRs) all the samples on one multi-sample TR may not
     necessarily.be in the same SDG.   In this instance,  the laboratory
     must make the appropriate number of photocopies of the TR so that a
     copy is submitted with each data package to which it applies.   In
     addition, in any instance where  samples from more than one multi-
     sample TR are in the same data package, the laboratory must submit a
     copy of the SDG cover sheet with copies of the TRs.

3.   High Concentration Volatiles Data

     a.    QC Summary

          (1)   Surrogate  Percent  Recovery  Summary  (Form II  HCV)

          (2)   Control Matrix  Spike Summary  (Form  III  HCV)

          (3)   Method Blank Summary (Form  IV HCV)

               (If more than a single  form is necessary,  forms must be
               arranged in chronological order by  date of analysis  of  the
               blank.)

          (4)   GC/MS Tuning and Mass Calibration  (Form V  HCV)

               BFB in chronological order; by instrument.
                              ' B-9                          Rev. 9/88

-------
b.   Sample Data

     Sample data shall be arranged in packets with the High
     Concentration Volatile Analysis Data Sheet (Form I HCV,
     including Form I HCV-TIC),  followed by the raw data for
     volatile samples.  These sample packets should then be placed
     in increasing EPA sample number order, considering both letters
     and numbers in ordering samples.  Phase units within a sample
     should be placed in order by phase unit suffix (i.e. AB123-11,
     AB123-12, etc.)

     (1)  TCL Results - High Concentration Volatile Analysis Data
          Sheet (Form I HCV).

          Tabulated results (identification and quantitation) of the.
          specified target compounds (Exhibit C).   The validation
          and release of these results is authorized by a specific,
          signed statement in the Case Narrative (reference C.I).
          In the event that the  Laboratory Manager cannot validate
          all data reported for  each sample,  the Laboratory Manager
          shall provide a detailed description of the problems
          associated with the sample in the Case Narrative.

          On Form I, the appropriate concentration units shall be
          mg/kg.  No other units are acceptable.  NOTE:  Report
          analytical results to  one significant figure if the value
          is less than 10;  to two significant figures above 10.

     (2)  Tentatively Identified Compounds (Form I TIC).

          This form must be included even if no compounds are found.
          If so, indicate this on the form by entering "0" in the
          field for "Number found."

          Form I HCV-TIC is the  tabulated list of the highest
          probable match for up  to 10 of the  nonsurrogate organic
          compounds not listed in Exhibit C (TCL),  including the CAS
          (Chemical Abstracts Registry)  number,  tentative
          identification and estimated concentration.   For
          estimating concentration, assume a response factor of  1,
          and estimate the concentration by comparison of the
          compound peak height or total  area count to the peak
          height or total area count of the nearest internal
          standard free of interferences on the reconstructed ion
          chromatogram.   NOTE:   The laboratory must be consistent
          (i.e., use peak height for all comparisons or use total
          area count for all comparisons).

     (3)  Reconstructed total ion chromatograms (RIG)  for each
          sample or sample extract.

          RICs must be normalized to the largest nonsolvent
          component, and must contain the following header
          information:
                          B-10
Rev. 9/88

-------
     o  EPA sample number
     o  Date and time of analysis

     o  GC/MS instrument ID

     o  Lab file ID

     Internal  standard and  surrogate spiking compounds are to
     be labeled with the names of compounds,  either directly
     out from  the peak,  or  on a print-out of retention times if
     retention times are printed over the peak.   If automated
     data system procedures are used for preliminary
     identification and/or  quantification of the Target
     Compound  List  (TCL) compounds,  the complete data system
     report must be included in all  sample data  packages,  in
     addition  to the reconstructed ion chromatogram.   The
     complete  data  system report shall include all of the
     information listed below.  For  laboratories which do  not
     use the automated data system procedures, a laboratory
     "raw data sheet,"  containing the following  information,
     must be included in the sample  data package in addition to
     the chromatogram.

     o  EPA sample number
     o  Date and time of analysis

     o  RT of identified TCL  compounds
     o  Ion used for quantitation with measured area
     o  Copy of area table from data system
     o  GC/MS instrument ID
     o  Lab file ID
(4)   For each  sample, by each  compound identified:

     (a)  Copies of raw  spectra and copies of
         background-subtracted mass spectra of  target
         compounds listed  in  Exhibit C  (TCL) that are
         identified in  the sample and corresponding
         background-subtracted TCL standard mass spectra.
         Spectra must be labeled with EPA sample number, lab
         file ID,  date  and time of analysis, and GC/MS
         instrument ID; compound names must be clearly marked
         on all spectra.

     (b)  Copies of mass spectra of nonsurrogate organic
         compounds not  listed in Exhibit C (TCL) (Tentatively
         Identified Compounds) with associated best-match
         spectra (three best matches), labeled as in  (4)(a)
         above.
                    B-ll
Rev. 9/88

-------
c.   Standards Data
                      •
     (1)   Initial  Calibration Data (Form VI  HCV)  -  in order by
          instrument,  if more than one  instrument used.

          (a)   HCV standard(s) reconstructed ion  chromatograms and
               quantisation reports  (or  legible facsimile)  for the
               initial  (five  point)  calibration,  labeled as  inb.(3)
               above.   Spectra are not required.

          (b)   All initial calibration data  must  be  included,
               regardless of  when  it was performed and for  which
               case.  When more  than one initial  calibration is
               performed, the data must  be put  in chronological
               order, by instrument.

     (2)   Continuing Calibration (Form  VII HCV)  - in order  by
          instrument,  if more than one  instrument used.

          (a)   HCV standard(s) reconstructed ion  chroraatograms and
               quantitation reports  (or  legible facsimile)  for all
               continuing (12 hour)  calibrations, labeled as in
               b.(3)  above.   Spectra are not required.

          (b)   When more than one  continuing calibration is
               performed, forms  must be  in chronological order,
               within fraction and instrument.

     (3)   Internal Standard Area Summary (Form  VIII  HCV)  -  in order
          by instrument, if more than one instrument used.

          When more than one  continuing calibration is performed,
          forms must be in chronological order, by  instrument.

d.   Raw QC Data

     (1)   BFB   (for each 12-hour period, for each GC/MS system
          utilized).

          (a)   Bar 'graph spectrum, labeled as in  b.(3) above.

          (b)   Mass listing,  labeled as  in b.(3)  above.

     (2)   Blank Data -  in chronological  order.  NOTE:  This order is
          different from that used for samples.

          (a)   Tabulated results (Form I  HCV).

          (b)   Tentatively Identified Compounds (Form I HCV-TIC)
               even if none found.

          (c)   Reconstructed  ion chromatogram(s) and quantitation
               report(s)  or legible  facsimile (GC/MS), labeled as in
               b.(3) above.

                         B-12                         Rev.  9/88

-------
               (d)  TCL spectra with lab generated standard, labeled as
                   In b.(4) above.  Data systems which are incapable of
                   dual display shall provide spectra in order:

                   o   Raw TCL compound spectra
                   o   Enhanced or background subtracted spectra
                   o   Laboratory  generated TCL  standard spectra
               (e)  GC/MS library search spectra for Tentatively
                   Identified Compounds (TIC), labeled as in b.(4)
                   above.

               (f)  Quantitation/Calculation of Tentatively Identified
                   Compound(s) (TIC) concentrations

          (3)   Control Matrix Spike Data

               (a)  Tabulated results (Form I HCV) of nonspiked TCL
                   compounds. Form I HCV-TIC not required.

               (b)  Reconstructed ion chromatogram(s) and quantitation
                   report(s) or legible facsimile (GC/MS), labeled as in
                   b.(4) above.  Spectra not required.
4.   Extractables Data
          QC  Summary

          (1)  Surrogate Percent Recovery Summary (Form II HCE)

          (2)  Control Matrix Spike Summary  (Form III HCE)

          (3)  Method Blank Summary (Form IV HCE)

              (If more than a single form is necessary, forms must be
              arranged in chronological order by date of analysis of the
              blank.)

          (4)  GC/MS Tuning and Mass Calibration (Form V HCE)

              DFTPP in chronological order; by instrument.

          Sample  Data

          Sample  data shall be arranged in packets with the High
          Concentration Extractable Analysis Data Sheet (Form I HCE,
          including Form I HCE-TIC), followed by the raw data for
          extractable samples.  These sample packets should then be
          placed  in increasing EPA sample number order, considering both
          letters and numbers in ordering samples.  Phase units within a
          sample  should be placed in order by phase unit suffix (i.e.,
          AB123-11, AB123-12, etc.).
                              B-13
Rev. 9/88

-------
(1)  TCL Results  -  High Concentration Extractable  Analysis  Data
     Sheet (Form  I  HCE-1,  HCE-2, HCE-3).

     Tabulated results  (identification and quantitation)  of the
     specified target compounds  (Exhibit  C).   The  validation
     and release  of these  results  is  authorized by a specific,
     signed statement in the  Case  Narrative  (reference  C.I).
     In the event that  the Laboratory Manager cannot validate
     all data reported  for each  sample, the Laboratory  Manager
     shall provide  a detailed description of  the problems
     associated with the sample  in the Case Narrative.

     On Form I, the appropriate  concentration units shall be
     mg/kg.   No other units are  acceptable.   NOTE:  Report
     analytical results to one significant figure  if the  value
     is less than 10; to two  significant  figures above  10.

(2)  Tentatively  Identified Compounds (Form I HCE-TIC).

     This form must be  included  even  if no compounds are  found.
     If so,  indicate this  on  the form by  entering  "0" in  the
     field for "Number  found".

     Form I HCE-TIC is  the tabulated  list of  the highest
     probable match for up to 20 of the nonsurrogate organic
     compounds not  listed  in  Exhibit  C (TCL),  including the CAS
     (Chemical Abstracts Registry)  number,  tentative-
     identification and estimated  concentration. For estimating
     concentration, assume a  response factor  of 1,  and  estimate
     the concentration  by  comparison  of the compound peak
     height or total area  count  to the peak height or total
     area count of  the  nearest internal standard free of
     interferences  on the  reconstructed ion chromatogram.
     NOTE:   The laboratory must  be consistent (i.e.,  use  peak
     height for all comparisons  or use total  area  count for all
     comparisons).

(3)  Reconstructed  total ion  chromatograms (RIG) for each
     sample, sample extract,  standard,  blank,  and  spiked
     sample.

     RICs must be normalized  to  the largest nonsolvent
     component, and must contain the  following header
     information:

     o    EPA sample number

     o    Date  and time of analysis
     o    GC/MS instrument ID
     o    Lab  file ID

     Internal standard  and surrogate  spiking  compounds  are  to
     be labeled with the names of  compounds,  either directly
     out from the peak,  or on a print-out of  retention  times If

                     B-14                         Rev. 9/88

-------
          retention times  are  printed  over  the peak.   If  automated
          data  system procedures  are used for preliminary
          identification and/or quantification of  the  Target
          Compound  List (TCL)  compounds, the complete  data  system
          report must be included in all sample data packages,  in
          addition  to the  reconstructed ion chromatogram.   The
          complete  data system report  shall include all of  the
          information listed below.  For laboratories  which do  not
          use the automated data  system procedures, a  laboratory
          "raw  data sheet," containing the  following information,
          must  be included in  the sample data package  in  addition to
          the chromatogram.

          o   EPA sample number

          o   Date and time of analysis

          o   RT of identified TCL compounds

          o   Ion used for quantitation with measured area

          o   Copy of area table from data system

          o   GC/MS instrument ID

          o   Lab file ID

     (4)   For each  sample, by  each compound identified:

          (a)   Copies of raw spectra and copies of
               background-subtracted mass spectra of target
               compounds .listed in Exhibit  C (TCL) that are
               identified  in the  sample and corresponding
               background-subtracted TCL standard mass spectra.
               Spectra must be labeled with EPA sample number,  lab
               file  ID, date and  time  of analysis, and GC/MS
               instrument  ID;  compound names must be clearly marked
               on all spectra.

          (b)   Copies of mass  spectra  of nonsurrogate  organic
               compounds not listed in Exhibit C (TCL) (Tentatively
               Identified  Compounds) with associated best-match
               spectra (three best matches), labeled as in  (4)(a)
               above.

c.   Standards  Data

     (1)   Initial Calibration  Data (Form VI HCE-1, HCE-2, HCE-3) -
          in order  by instrument,  if more than one instrument used,

          (a)   Extractables standard(s) reconstructed  ion
               chroraatograms and  quantitation reports  (or legible
               facsimile)  for the  initial (three point) calibration,
               labeled as  in b.(3) above.    Spectra are not  required.

          (b)  All  initial calibration data must be included,
               regardless of when  it was performed and for  which


                         B-15                         Rev.  9/88

-------
              case.  When more  than  one  initial  calibration is
              performed, the data must be put  in chronological
              order, by  instrument.

     (2)  Continuing  Calibration (Form VII HCE-1, HCE-2,  HCE-3)  -  in
          order  by instrument,  if more than  one instrument used.

          (a)  Extractable standard(s) reconstructed  ion
              chromatograms and quantitation reports  (or legible
              facsimile) for all continuing (12  hour) calibrations,
              labeled as in b.(3) above.  Spectra are not required,

          (b)  When more  than one continuing calibration  is
              performed, forms  must  be in chronological  order, by
              instrument.

     (3)  Internal Standard Area Summary  (Form  VIII HCE-1,  HCE-2)  -
          in order by instrument, if  more than  one instrument used.

          When more than  one continuing calibration is performed,
          forms  must  be in chronological  order  by instrument.

     (4)  GPC Calibration (Form  IX HCE) - in order by calibration
          date.

          (a)  Copies of UV traces for each GPC calibration
              performed  labeled with date and  time of the
              calibration, and  labeling  each peak in  the
              calibration standard with  the name of  the  compound
              and its retention time.             ••—-

d.   Raw QC Data

     (1)  DFTPP  (for  each 12-hour period, for each GC/MS  system
          utilized)

          (a)  Bar graph  spectrum, labeled as in  b.(3) above.

          (b)  Mass listing, labeled  as in b.(3)  above.   - -

     (2)  Blank  Data  - in chronological order.  NOTE:  This  order  is
          different from  that used for samples.

          (a)  Tabulated  results  (Form I  HCE-1, HCE-2, HCE-3)
                                     •
          (b)  Tentatively Identified Compounds (Form I HCE-TIC) -
              even if none found.

          (c)  Reconstructed ion  chromatogram(s)  and quantitation
              report(s) or legible facsimile (GC/MS), labeled as in
              b.(3) above.
                         B-16                         Rev. 9/88

-------
          (d)  TCL spectra with lab generated standard,  labeled as
               in b.(4)  above.   Data systems  which are incapable of
               dual display shall provide  spectra in order:

               o  Raw TCL compound spectra

               o  Enhanced or background subtracted spectra

               o  Laboratory generated TCL standard spectra

          (e)  GC/MS library search spectra for  Tentatively
               Identified Compounds (TIC), labeled as  in b.(4)
               above.

          (f)  Quantitation/Calculation of Tentatively Identified
               Compound(s)  (TIC)  concentrations.

     (3)  Control Matrix Spike  Data

          (a)  Tabulated results  (Form I)  of  nonspiked TCL
               compounds.  Form  1  HCE-TIC not  required.

          (b)  Reconstructed ion  chromatogram(s)  and quantitation
               report(s)  or  legible facsimile (GC/MS),  labeled  as in
               b.(3) above.   Spectra  not required.

Aroclor/Toxaphene Data .

a.   QC Summary

     (1)  High Concentration Aroclor Surrogate Recovery Summary
          (Form II HCA)

     (2)  High Concentration Aroclor Control  Matrix Spike Recovery
          Summary (Form III HCA)

     (3)  High Concentration Aroclor Method Blank Summary (Form IV
          HCA-1)

          (If more than a single form is necessary, forms must  be
          arranged in chronological order  by  date of analysis of the
          blank.)

     (4)  High Concentration Aroclor Instrument  Blank Summary (Form
          IV HCA-2)

          (If more than a single  form is necessary, forms must  be
          arranged in chronological order  by  date of analysis of the
          blank.)

b.   Sample Data

     Sample data shall be arranged in packets with the High
     Concentration Aroclor Analysis Data Sheet (Form I HCA),
     followed by the raw data for samples.  These sample packets


                         B-17                         Rev. 9/88

-------
     should then be placed in increasing EPA sample  number order,
     considering both letters and numbers in ordering samples.

     (1)  TCL Results - High  Concentration Aroclor Analysis Data
         Sheet  (Form I HCA).

         Tabulated results  (identification  and quantitation) of  the
         specified target compounds  (Exhibit C).  The validation
         and release of  these results  is authorized by  a specific,
         signed statement in the Case  Narrative  (reference C.I).
         In the event that  the Laboratory Manager cannot validate
         all data  reported for each  sample,  the  Laboratory Manager
         shall  provide a detailed description of the problems
         associated with the sample  in the  Case  Narrative.

         On Form I,  the  appropriate  concentration units shall  be
         mg/kg. No other units are  acceptable.

         NOTE:  Report analytical results to two  significant figures
         for all Aroclor/Toxaphene samples.

     (2)  Copies of Aroclor/Toxaphene analysis chroraatograms.

         All chromatograms  must be labeled  with  the following
         information:

         o  EPA sample number
         o  Volume  injected  (ul)

         o  Date and time of  injection

         o  GC column identification  (by stationary phase)
         o  GC instrument identification

         o  Positively identified compounds must be labeled with
             the names of compounds,  either directly out from the
             peak,  or on a print-out of retention times if
             retention times are printed over the peak.

     (3)  Copies of Aroclor/Toxaphene analysis chromatograms  from
         second GC column confirmation.   Chromatograms  to  be
         labeled as in (2) above.

     (4)  Manual work sheets.

c.   Standards  Data

     (1)  Form VI HCA-1 and HCA-2 - High  Concentration Initial
         Calibration of  Multicomponent Analytes  (all GC columns)

     (2)  Form VII  HCA -  High Concentration  Continuing Calibration
         of Multicomponent Analytes  (all GC  columns)

     (3)  Form VIII  HCA -  High Concentration  Analytical  Sequence
         (all GC columns)

                         B-18                         Rev. 9/88

-------
     (4)   Form IX HCA -  High Concentration  Single  Component
          Pesticide  Retention Times  (all GC columns)

     (5)   Form X HCA - High  Concentration Aroclor  Identification
          Summary (only  required  for positive  results)

     (6)   Form XI HCA -  High Concentration  Aroclor Diol  Cartridge
          Check (all lot numbers  used)

     (7)   Aroclor/Toxaphene  standard chromatograms and data system
          printouts  for  all  standards to include:

          o  All Aroclors and Toxaphene

          o  All quantitation standards

          o  Diol cartridge check standard for each lot of
             cartridges used
          o  A copy of the computer reproduction or strip chart
             recorder output covering the 100 fold range

          (a)  All chromatograms  are  required  to have the following:

              o   Label all standard peaks  for all individual
                  compounds either directly out from the peak or on
                  the printout of retention times  if retention times
                  are printed over the peak.
              o   Label the chromatogram for multicomponent
                  standards, (i.e.,  Aroclor 1242,  Toxaphene)

              o   List total ng injected for each standard.
              o   A printout of retention times and corresponding
                  peak areas must accompany each chromatogram.

              o   Date and time of injection.
              o   GC column identification (by stationary phase).

              o   GC instrument identification.

d.   Raw QC Data

     (1)   Method Blank Data  -  in  chronological order.  NOTE:   This
          order is different from that used for samples.

          (a)  Tabulated results  (Form I HCA).

          (b)  Chromatogram(s) and data system printout(s)  (GC)  for
              each  GC column and instrument used  for analysis,
              labeled as  in b.(2) above.

     (2)   Instrument Blank Data  - in chronological order.   NOTE:
          This order is  different from that used for samples.

          (a)  Tabulated results  (Form I HCA).

          (b)  Chromatogram(s) and data system printout(s)  (GC)  for
              each  GC column and instrument used  for analysis,
              labeled as  in b.(2) above.

                          B-19                         Rev. 4/89

-------
               (3)  Control Matrix Spike Data

                   (a)  Tabulated results (Form I HCA) of nonspike TCL
                        compounds.

                   (b)  Chromatogram(s) and data system printout(s) (GC),
                        labeled as in b.(2) above-.

F.   GC/MS Tapes

     The Contractor must store all raw and processed GC/MS data on magnetic
     tape, in appropriate instrument manufacturer's format.   This tape must
     include data for samples, blanks, control matrix spikes, initial
     calibrations,  continuing calibrations,  BFB and DFTPP, as well as all
     laboratory-generated spectral libraries and quantitation reports
     required to generate the data package.   The Contractor shall maintain a
     written reference logbook of tape files to EPA sample number,
     calibration data, standards, blanks,  and control matrix spikes.   The
     logbook should include EPA sample numbers and standard and blank ID's,
     identified by Case and Sample Delivery Group.

     The Contractor is required to retain the GC/MS tapes for 365 days after
     data submission.   During that time,  the Contractor shall submit  tapes
     and associated logbook pages within seven days after receipt of  a
     written request from the Project Officer or the Sample Management
     Office.

G.   Extracts                       •

     The Contractor shall preserve sample extracts at 4°C (±2"C) in
     bottles/vials with teflon-lined septa.   Extract bottles/vials shall be
     labeled with EPA sample number,  Case number and Sample Delivery  Group
     (SDG) number.   A logbook of stored extracts shall be maintained,  listing
     EPA sample numbers and associated Case and SDG numbers.

     The Contractor is required to retain extracts for 365 days following
     data submission.   During that time,  the Contractor shall submit  extracts
     and associated logbook pages within seven days following receipt of a
     written request from the Project Officer or the Sample  Management
     Office.

H.   Complete Case File Purge

     (Formerly, Document Control and Chain-of-Custody Package).

     The complete case file purge includes all laboratory records received or
     generated for a specific Case  that have not been previously submitted to
     EPA as a deliverable.  These items include but are not limited to:
     sample tags, custody records,  sample tracking records,  analysts  logbook
     pages, bench sheets, chromatographic charts,  computer printouts,  raw
     data summaries,  instrument logbook pages,  correspondence,  and the
     document inventory (see Exhibit F).
                                   B-20                         Rev.  9/88

-------
                                  SECTION III


                            FORM INSTRUCTION GUIDE


This section includes specific instructions for the completion of all
required forms.  Each of the forms is specific to a given fraction (volatile,
or extractable).   The contractor shall submit only those forms pertaining to
the fractions analyzed for a given sample or samples.  For instance, if a
sample is scheduled for high concentration volatile analysis only, provide
only HCV forms.  There are three pages relating to the extractable fraction
for Forms I, VI,  VII, and VIII.  Whenever extractables are analyzed and one
of the above named forms is required, all three pages {HCE-1, HCE-2, and HCE-
3) must be submitted.  In addition to the forms for high concentration
volatiles and high concentration extractable, forms are provided for the
Aroclor-specific GC/EC analysis of high concentration samples (HCA).  These
instructions are arranged in the following order:

     A.  General  Information and Header  Information

     B.  Organic  Analysis Data  Sheets  (Form  I, All Fractions)
     C.  Surrogate Recovery (Form  II,  All Fractions)

     D.  Control  Matrix  Spike Recovery (Form III, All  Fractions)
     E,  Method Blank Summary (Form IV,  All  Fractions)
     F.  GC/MS  Tuning and Mass  Calibration  (Form V HCV,  Form V HCE)
     G.  Initial  Calibration Data'(Form  VI HCV, Form VI  HCE, Form VI HCA)
     H.  Continuing Calibration Data  (Form VII HCV, Form VII HCE, Form VII
         HCA)
     I.  Internal Standard Area Summary  (Form VIII HCV,  Form VIII HCE) and
         Analytical Sequence (Form VIII  HCA)
     J.  GPC Calibration (Form  IX HCE) and Pesticide Retention Times (Form IX
         HCA)
     K.  Aroclor  Identification Summary  (Form X HCA)

     L.  Diol Cartridge Check (Form XI HCA)
                                   B-21
Rev. 9/88

-------
A.
General Informationand Header Information
Values must be reported on the forms according to the individual form
instructions in this Section.  For example, results for concentrations of HCV
TCL compounds must be reported to two significant figures if the value is
greater than or equal to 10.

All characters which appear on the data reporting forms presented in the
contract (Exhibit B, Section IV) must be reproduced by the contractor when
submitting data, and the format of the forms submitted must be identical to
that shown in the contract.  No information may be added, deleted, or moved
from its specified position without prior written approval of the EPA Project
Officer.  The names of the various fields and compounds (i.e., "Lab .Code,"
"Chloromethane") must appear as they do on the forms in the contract.  For
items appearing on the uncompleted forms (Section IV), the use of uppercase
and lowercase letters is optional.

Do not remove the underscores or vertical bar characters that delineate
"boxes" on the forms.  The only exception would be those underscores at the
bottom of a "box" that are intended as a data entry line (for instance, see
Form 2HA,. line 30.  If data must be entered on line 30, it will replace the
underscores).

Six pieces of information are common to the header sections of each data
reporting form.  They are:  Lab Name, Contract, Lab Code, Case No., SAS No.,
and SDG No.  This information must be entered on every form and must match on
every form.

The "Lab Name" shall be the name- chosen by the contractor to identify the
laboratory.

The "Lab Code" is an alphabetical abbreviation of up to 6 letters, assigned
by EPA, to identify the laboratory and aid in data processing.  This lab code
shall be assigned by EPA at the time a contract is awarded, and shall not be
modified by the contractor, except at the direction of EPA.

The "Case No." is the EPA-assigned Case number (up to 5 digits) associated
with the sample, and reported on the Traffic Report.

The "Contract" is the number of the EPA contract under which the analyses
were performed.

The "SDG No." is the Sample Delivery Group number.   The Sample Delivery Group
(SDG) number is the EPA Sample Number of the first sample received in the
SDG.  When several samples are received together in the first SDG shipment,
the SDG number shall be the lowest sample number (considering both alpha and
numeric designations) in the first group of samples received under the SDG.

The "SAS No." is the EPA-assigned number for analyses performed under Special
Analytical Services.  If samples are to be analyzed under SAS only, and
reported on these forms, then enter SAS No.,  and leave Case No.  blank.  If
samples are analyzed according to the "Routine Analytical Services" (IFB)
protocols and have additional "SAS" requirements,  list both Case No.  and SAS
                                   B-22
                                                          Rev. 9/88

-------
No. on all forms.  If the analyses have no SAS requirements, leave "SAS No."
blank.  NOTE:  Some samples in an SDG may have a SAS No. while others do not.

The other information common to most of the forms is the "EPA Sample No.".
This number appears either in the upper right-hand corner of the form, or as
the left column of a table summarizing data from a number of samples. When
"EPA Sample No." is entered into the triple-spaced box in the upper right-
hand corner of Form I, it should be entered on the middle line of the three
lines that comprise the box.

All, samples, control matrix spikes, blanks and standards shall be identified
with an EPA Sample Number.  Because a sample may comprise a number of single-
phase units, the use of sample number suffixes is necessary to differentiate
between phase units.  The following system of suffixes must be used:

     XXXXX      -  EPA sample number

     XXXXX-11   -  first phase unit

     XXXXX-12   =  second phase unit

     through

     XXXXX-19   -  ninth phase unit

     XXXXX-YYRE -  reanalysis of phase unit YY

     XXXXX-YYDL -  phase unit YY analyzed at a secondary dilution.

     CMS        -  Control Matrix Spike

     Note:  If more than nine phases are separated from a single sample,
     contact SMO for instructions on numbering the phases.

Use of the suffixes "RE" and/or "DL" assumes that data are also being
submitted for an analysis of a phase unit that represents the "original"
analysis of the phase unit.  If only one set of data are being submitted, do
not use the suffix "RE" or "DL".

For blanks and standards, the following identification scheme must be used as
the "EPA Sample No."

     1.   Volatile  blanks  shall be  identified  as  VBLK##.

     2.   Extractable blanks shall  be  identified  as  EBLK##.

     3.   Aroclor method blanks shall be  identified  as ABLK##.

     4.   Aroclor instrument blanks  shall be identified  as IBLK##.

         The  "EPA  Sample No." must  be unique  for each blank within  an SDG.
         Within a  fraction, a laboratory must achieve this by replacing  the
         two-character "##" terminator of the identifier with one or  two


                                   B-23                          Rev.  9/88

-------
          characters  or numbers,  or  a combination  of both.   For example,
          possible  identifiers  for volatile blanks would be  VBLK1,  VBLK2,
          VBLK10, VBLKAB,  etc.

     5.    Volatile  and extractable standards  shall be  identified as:

          ,  FSTD###,  where:

            F - fraction (V for volatiles; E for extractable).

            ### - the concentration in ug/L of volatile standards (i.e.,  20,
            50, 100,  150,  200)

            or

            the amount injected in ng for extractable standards (i.e. 50,  80,
            160).   For PCB standards, the amounts  injected would be 10, 30,
            50 or 20, 60,  100 ng.  If PCB standards  are coinjected in
            different amounts,  use Che lower amount  (i.e., ESTD10 for
            coinjection of 10 and 20 ng of PCBs)

     6.    Aroclor standards shall be identified as specified in the
          instructions for Form VIII.

Several other pieces  of information are common to  many of the Data Reporting
Forms.  These include:  Phase type,  Phase weight,  and Lab Sample ID.  '

For the purposes of this contract, there are only three possible phase types.
They are:

      Solids
      Water Immiscible Liquids
      Water Miscible Liquids

For "Phase Type",  enter "SOLID" for solid phase units.  Enter "WIL" for water
immiscible liquids, and enter "WML" for water miscible liquids.  Water
samples are considered as water miscible liquids,  and would be entered as
"WML", not "WATER".

For "Phase weight"  enter the number of grams of phase unit used.

"Lab Sample ID" is  an optional laboratory-generated internal identifier.

"Lab File ID" is the laboratory-generated name of the GC/MS data system file
containing information pertaining to a particular  analysis.   Up to 14
alpha-numeric characters may be used here.

Forms II, IV, V, and VIII contain a field labeled "page _ of _" in the bottom
left-hand corner.   If the number of entries  required on any of these  forms
exceeds the available space, continue entries on another copy of the  same
fraction-specific form, duplicating all header information.  If a second page
is required, number them consecutively, as "page 1 of 2" and "page 2  of 2".
If a second page is not required, number the page  "page 1 of 1."


                                   B-24                         Rev.  9/88

-------
For rounding off numbers to the appropriate level of precision, observe the
following common rules.  If the figure following those to be retained is less
than 5, drop it (round down).  If the figure is greater than 5, drop it and
increase the last digit to be retained by 1 (round up).  If the figure
following the last digit to be retained equals 5, round up if the digit to be
retained is odd, and round down if that digit is even.

B.   Organic Analysis Data Sheet (Form I)

     1.   Form  I HCV,  Form I  HCE-1,  Form  I  HCE-2,  Form I  HCE-3,  Form I  HCA

            This form is used for tabulating and reporting sample analysis
            results for Target Compound List (TCL) compounds.  If all
            fractions are not requested to be analyzed, only the pages
            specifically required must be submitted.  If HCV analysis only is
            requested, Form I HCV and Form I HCV TIC must be submitted.  If
            the Aroclor-specific analysis is the only analysis requested,
            only Form I HCA must be submitted for that sample.

            Complete the header information on each page of Form I required,
            according to the instructions in part A. and as follows:

            Enter pH for Extractables and Aroclors, reported to 0.1 pH units.

            "Date Received" is the date of sample receipt at.the laboratory,
            as noted on the Traffic. Report (i.e., the VTSR).  It should be
            entered as MM/DD/YY.

            "Date Separated", "Date Extracted" and "Date Analyzed" should be
            entered in a similar fashion.  For Aroclor-specific samples, the
            date of analysis should be the date of the first GC analysis
            performed.

            For all fractions, enter the final volume in milliliters of the
            sample extract under "Final Extract Volume".   Report volume to
            one tenth of a milliliter (i.e. 10.0, not 10).

            For volatiles, enter the number of microliters of the extract
            that was added to the reagent water under "Aliquot Volume".

            For extractable,  enter the number of microliters of the extract
            that is injected under "Injection Volume".

            For all three fractions,  enter under "Conversion Factor" the
            value  used to convert your raw data into the  concentration values
            in  mg/Kg reported on Form I for all detected  compounds.   Note:
            this conversion factor must incorporate the phase weight, the
            final  extract volume,  the injection or aliquot volume,  the use of
            GPC clean-up procedures,  and  any dilution of  the extract or
            sample.that is required.   The units associated with this factor
            will vary,  depending on the units given in the raw data.
            Therefore,  no units  for the conversion factor are to be  reported
            on  Form I.
                                   B-25                          Rev.  9/88

-------
u -
J -
B -
If samples for Aroclor/Toxaphene analysis were subjected to
sulfur clean-up procedures, enter "Y" under "Sulfur Clean-up
(Y/N)."  If no sulfur clean-up was performed enter "N".

For positively identified TCL compounds, the contractor shall
report the concentrations detected as uncorrected for blank
contaminants.

For volatile and extractable results, report analytical results
to one significant figure if the value is less than 10, and two
significant figures above 10.

Report all Aroclor-specific results to two significant figures.

The concentration units are mg/Kg for all phase types and
fractions.

If the result is a value greater than or equal to the
quantitation limit, report the value.

Under the column labeled "Q" for qualifier, flag each result with
the specific Data Reporting Qualifiers listed below.  The
Contractor is encouraged to use additional flags or footnotes.
The definition of such flags must be explicit and must be
included in the Case Narrative.

For reporting results to the USEPA," the following contract
specific qualifiers are to be used.  The eight qualifiers defined
below are not subject to modification by the laboratory.  Up to
five qualifiers may be reported on Form I for each compound.

The eight EPA-defined qualifiers to be used are as follows:

Indicates compound was analyzed for but not detected.  The sample
quantitation limit must be corrected for'dilution.  For example,
20 U for phenol if the sample final volume is the
protocol-specified final volume.  If a 1 to 10 dilution of
extract is necessary, the reported limit is 200 U.
Indicates an estimated value.  This flag is used either when
estimating a concentration for tentatively identified compounds
where a 1:1 response is assumed, or when the mass spectral or
GC/EC data indicate the presence of a compound that meets the
identification criteria but the result is less than the sample
quantitation limit but greater than zero.  For example, if the
sample quantitation limit is 10 mg/Kg, but a concentration of 3
mg/Kg is calculated, report it as 3J.  The sample quantitation
limit must be adjusted for dilution as discussed for the U flag.

This flag is used when the analyte is found in the associated
blank as well as in the sample.  It indicates possible/probable
blank contamination and warns the data user to take appropriate
action.  This flag must be used for a TIC as well as for a
positively identified TCL compound.
                             B-26
                                                    Rev. 9/88

-------
      E -   This flag identifies compounds whose concentrations exceed the
            calibration range of the GC/MS instrument for that specific
            analysis.  This flag will not apply to Aroclors analyzed by GC/EC
            methods.  If one or more compounds have a response greater than
            full scale, the extract must be diluted and re-analyzed according
            to the specifications in Exhibit D. All such compounds with a
            response greater than full scale should have the concentration
            flagged with an "E" on the Form I for the original analysis.  If
            the dilution of the extract causes any compounds identified in
            the first analysis to be below the calibration range in the
            second analysis, then the results of both analyses shall be
            reported on separate Forms I.   The Form I for the diluted sample
            shall have the "DL" suffix appended to the sample number.

      D -   This flag identifies all compounds identified in an analysis at a
            secondary dilution factor.  If a sample or extract is re-analyzed
            at a higher dilution factor, as in the "E" flag above, the "DL"
            suffix is appended to the sample number on the Form I for the
            diluted sample, and all concentration values reported on that
            Form I are flagged with the "D" flag.

      A -   This flag indicates that a TIC is a suspected aldol-condensation
            product.

      N -   This flag identifies Aroclor or Toxaphene compounds where one or
            more of the peaks used for quantitation are more than two times
            the width of the corresponding peaks in the highest concentration
            calibration standard.  It indicates an uncertainty in the
            quantitation for the compound other than those discussed under
            the "J" flag.
      X -   Other specific flags and footnotes may be required to properly
            define the results.  In order to limit the number of laboratory-
            defined flags and not use such flags as may be part of the
            Agency's data review processes, the laboratory-defined flags are
            restricted to the three letters "X", "Y", and "Z".  If used, they
            must be fully described and such description attached to the
            Sample Data Summary Package and the Case Narrative.  If more than
            one is required, use "Y" and "Z", as needed.  If more than five
            qualifiers are required for a sample result, use the "X" flag to
            combine several flags, as needed.  For instance, the "X" flag
            might combine the "A", "B", and "D" flags for some samples.

     The combination of flags "BU" or "UB" is expressly prohibited.  Blank
     contaminants are flagged "B" only when they are also detected in the
     sample.

     If analyses at two different dilution factors are required (see Exhibit
     D),  follow the data reporting instructions given in Exhibit D and with
     the "D"  and "E" flags above.

2.   Form I HCV-TIC and Form I HCE-TIC

     Fill in all header information as above.
                                   B-27
Rev. 9/88

-------
     Report Tentatively Identified Compounds (TIC) including CAS number,
     compound name, retention time, and the estimated concentration (criteria
     for reporting TICs are given in Exhibit D, Section IV).  Retention time
     must be reported in minutes and decimal minutes, not seconds or
     minutes:seconds.

     If in the opinion of the mass spectral interpretation specialist, no
     valid tentative identification can be made,  the compound shall be
     reported as unknown.

     Include a Form I HCV-TIC or HCE-TIC for every volatile and extractable
     fraction of every sample and method blank analyzed, even if no TICs are
     found.  Total the number of TICs found, including aldol-condensation
     products (but see below),  and enter this number in the "Number TICs
     found." If none were found, enter "0" (zero).
     If the name of a compound exceeds the 28 spaces in tha TIC column,
     truncate the name to 28 characters.  .If the compound 'is an unknown,
     restrict description to no more than 28 characters (i.e., unknown
     hydrocarbon, etc.).                                  !

     Peaks that are suspected as aldol-condensation reaction products (i.e.,
     4-methyl-4-hydroxy-2-pentanone and 4-methyl-3-pentene-2-one) shall be
     summarized on this form, flagged "A", and included in the total "Number
     TICs found," but not counted as part of the 20 most intense non-TCL
     extractable compounds to be searched.  NOTE:  The nam
-------
     stationary phase in the fields labelled "GC Column ID (1)" and "GC
     Column ID (2)".

     If an interference prevents the quantitation of one of the two Aroclor
     surrogates, enter that value as "0", flag it in column under the "#"
     symbol, and describe the problem in the Case Narrative.

     Number all pages as described in part A.

D.   Control Matrix Spike Recovery (Form III)

     This form is used to report the results of the analyses of the control
     matrix spike.

     Complete the header information as instructed in Fart A.

     In the box on Form III, under "SPIKE ADDED", enter the calculated
     concentration in mg/Kg that results from adding each spiked compound to
     the aliquot chosen for the control matrix spike (CMS).  For instance,
     for extractable compounds, if 100 ug of spike are added to 1 g of
     sample, the resulting concentration is 100 mg/Kg.  Under "CMS
     CONCENTRATION", enter the actual concentration of each spike compound
     detected in the control matrix spike aliquot.  Calculate the percent
     recovery of each spike compound in the control matrix spike to the
     nearest whole percent,  according to Exhibit E,  and enter under "CMS %
     REG".  Flag all percent recoveries outside the QC limits with an
     asterisk (*).  The asterisk must be placed in the last space of the
     percent recovery column,  under the "#" symbol.                     . .__...

     On Form III HCA, there are no QC limits on recovery at this time.

     For volatiles and extractables,  summarize the values outside the QC
     limits at the bottom .of the page.   No further action is required by the
     laboratory.   Performance-based QC limits will be generated and updated
     from recovery data.

E.   Method Blank Summary  (Form IV)

     This form summarizes the samples associated with each method blank
     analysis.  A copy of the appropriate Form IV is required for each blank.

     Complete the header information on Form IV as described in Part A.

     For volatile and extractable blanks, enter the "Instrument ID", "Date
     Analyzed",  and "Time Analyzed".   The "Time Analyzed" shall be in
     military time.

     Forrextractable blanks and Aroclor method blanks, enter the date of
     extraction of the blank.

     If some of the samples associated with a given method blank are
     subjected to sulfur clean-up procedures and some samples are not,  then
     two method blanks are required,  one subjected to sulfur clean-up and one
     not.  (See Exhibit D for suggestions regarding the aliquotting of method

                                   B-29                         Rev. 9/88

-------
F.
blank extracts).  If the method blank has been subjected to sulfur
clean-up, then enter "Y" under "Sulfur Clean-up (Y/N)," and list in the
table only those samples associated with that blank that have also
undergone sulfur clean-up.  Complete a separate Form IV HCA-1 for the
aliquot of the method that did not undergo sulfur clean-up, listing the
associated samples that also did not undergo sulfur clean-up.  NOTE:
The aliquotting of the method blank will require that different blank
identifiers be used in place of the EPA Sample Number for each blank on
all deliverables.  (See Section A above).

Aroclor/Toxaphene contaminants must meet the identification criteria in
Exhibit D ARO, which requires analysis of the blank on two different GC
Columns.  Therefore, enter the date, time and instrument ID of both
analyses on the pesticide method blank summary.  The information on the
two analyses is differentiated as Date Analyzed (1), Date Analyzed (2),
etc.  If the analyses were run simultaneously, the order of reporting is
not important, but must be consistent with the information reported on
Form X.  Otherwise (1) shall be the first analysis, and (2) the second.
Identify both GC columns by stationary phase under "GC Column ID".

Aroclor/Toxaphene analyses also require the analysis of an instrument
blank at specified points in the analytical sequence.  For each
instrument blank that is associated with phase units in a data package,
complete a copy of Form .IV HCA-2, and summarize the samples associated
with that instrument blank, as described below.  NOTE:  The samples
associated with an instrument blank may not be the same as those
associated with a method blank.  Samples associated with an instrument
blank are those that were analyzed in the 12-hour period prior to the
instrument blank in question, and after the previous acceptable
instrument blank.

For all three fractions, as appropriate, summarize the samples
associated with a given method blank in the table below the header,
entering EPA Sample Number, and Lab Sample ID.  For volatiles, enter the
Lab File ID and time of analysis of each sample.  For extractables,
enter Lab File ID and the date and time of analysis of each sample.

Number all pages as described in part A.

GC/MS Tuning and Mass Calibration (Form V)

This form is used to report the results of GC/MS tuning for volatiles
and extractable, and to summarize the date and time of analysis of
samples, standards, blanks, and control matrix spikes associated with
each GC/MS tune.

Complete the header information as in part A.  Enter the "Lab File ID"
for the injection containing the GC/MS tuning compound (BFB for
volatiles, DFTPP for extractable).  Enter the "Instrument ID".  Enter
the date and time of injection of the tuning compound.   Enter time as
military time.

For each ion listed on the form, enter the percent relative abundance in
the right-hand column.  Report relative abundances to the number of
                                   B-30
                                                           Rev. 9/88

-------
 significant  figures  given  for  each  ion  in  the  ion abundance  criteria
 column.

 All  relative abundances  must be  reported as a  number.   If zero,  enter
 "0",  not  a dash  or other non-numeric character.  Where  parentheses
 appear, compute  the  percentage of the ion  abundance  of  the mass  given  in
 the  appropriate  footnote,  and  enter that value in the parentheses.

 In the  lower half of the form, list all samples, standards,  blanks,  and
 control matrix spikes analyzed under that  tune in chronological  order.
 by time of analysis  (in  military time).  Refer to part  A. for  specific
 instructions for identifying standards  and blanks.   Enter "EPA Sample
 No.", "Lab Sample ID", "Lab File ID", "Date Analyzed",  and "Time
 Analyzed" for all standards, samples, blanks,  and contro.l matrix spikes.

 The  GC/MS tune expires twelve  hours from the time of injection of the
 tuning  compound  (BFB or  DFTPP) listed at the top of  the form.  In order
 to meet the  tuning requirements, a  sample, standard, blank,  or  control
 matrix  spike must be injected  within twelve hours of the injection of
 the  tuning compound.

 Number  all pages as  described  in part A.

 Initial Calibration  Data (Form VI)

 After a GC/MS system has undergone  an initial  calibration at the
 specific concentration levels  described in Exhibit E, and after  all
 initial calibration  criteria have been met, the laboratory must  complete
 and  submit a Form VI  for each  volatile or  extractable TCL initial
 calibration  performed which is relevant to the  samples, blanks,  or
 control matrix spikes in the SDG, regardless of when that calibration
 was  performed.   A five-point initial calibration is  required for
 volatiles.   A three-point  calibration is required for extractables.

 Complete all  header  information  as  in part A.   Enter the "Case No."  and
 "SDG No." for the current  data package,  regardless of the original Case
 for which the  initial calibration was performed.  Enter "Instrument  ID"
 and  the date(s)  of the calibration.   If the calendar date changes during
 the calibration  procedure,  the inclusive dates should be given on Form
VI.  Enter the "Lab File ID" for each of the calibration standards
 injected.   Complete the  response factor data for the calibration  points,
 and  then calculate and report the average relative response  factor (RRF)
 for all TCL  and  surrogate compounds.  The laboratory must report  the
%RSD for all  compounds. All CCC compounds must have a %RSD of less than
or equal to  30.0 percent. All HCV SPCC compounds must have a minimum
average relative response factor (RRF)  of 0.300 (0.250 for Bromoform).
All extractable  SPCC compounds must  have a minimum average relative
response factor  (RRF) of 0.050.
       %RSD -
                SD
x 100
                              B-31
                                                           Rev.  9/88

-------
       where:   %RSD
Relative Standard Deviation
       SD  =  Standard Deviation of initial response factors (per
       compound)
               n-1
        IT  -  mean of initial response factors (per compound)

The PCBs listed on Form VI HCE-3 are injected in different amounts than
the other extractable standards, and the octa-decachlorinated biphenyls
are injected at twice the amount as the lower levels of chlorination.
Enter the two sets of Lab File IDs for the injections of the PCBs in the
appropriate spaces on the lower half of this form.  If the FCBs are
injected along with the other standards, enter the same Lab File IDs as
at the top of the form.  If the PCBs are all injected in one standard
regardless of level of chlorination, but separate from the other
extractable standards ,  enter one set of Lai; File IDs for the extractable
standards, and enter a second set of Lab File IDs for the PCBs, using
the same Lab File ID for RRF10 as for RRF20, etc.

The initial calibration of the Aroclors and Toxaphene is reported on
Form VI HCA-1 and HCA-2.  Complete all header information as in part A.
For each GC column used for Aroclor/Toxaphene analyses, complete a copy
of Form VI HCA-1 and HCA-2, identifying the stationary phase of the GC
Column under "GC Column ID" .  Each analyte requires a three-point
calibration, therefore, for each GC Column, there will be three pages of
Form VI HCA-1.  The "Date(s) Analyzed" at the top of the form must be
the inclusive dates of analysis for all 24 of the standards in the
initial calibration.

For each injection of each standard, enter the date and time of
injection, and the amount injected in nanograms .

The calibration of these raulticomponent analytes requires the use of at
least three chromatographic peaks.  Two additional peaks may be reported
as well to allow for coeluting interferences.  For each peak used, enter
the retention time in minutes and decimal minutes (not minutes and
seconds), and the calibration factor for each peak.

Calculate a mean of the three retention times for each peak of each
analyte, and enter under "MEAN RT" .   Calculate a mean deviation at the
three retentions for each peak, according to the formula in Ex. D. ARO,
and report this value under "MD RT" .

Calculate a mean and relative standard deviation of the three
calibration factors for each peak in each analyte according to Ex. D
ARO, and enter them on Form VI HCA-2 under "Mean Cal.  Factor" and
"%RSD".
                              B-32
                                   Rev. 4/89

-------
     If the mean and %RSD do not meet the specifications in Ex. D. ARO, the
     laboratory may use a regression line through all three calibration
     points, or a two segment regression line (see Ex. D. ARO).  If the mean
     and %RSD meet the specifications in Ex. D.  ARO, leave blank the
     regression coefficients and intercept fields.

     If a single correlation line is used, enter the mean, %RSD, and the
     regression coefficient "r (1)" and the intercept "INTCPT (1)".

     If a two segment regression line is used, enter both correlation
     coefficients and intercepts "r (l)"and "r (2)", etc.  Also enter the
     mean RT and %RSD.

H.   Continuing Calibration Data (Form VII)

     1.    Form VII HCV,  HCE-1, HCE-2,  and  HCE-3

          The  Continuing Calibration Data  Form is  used to verify the
          calibration of the GC/MS  system  by  the analysis of  specific
          calibration standards.  A Continuing Calibration Data Form is
          required for each twelve  (12)  hour  time  period for  both volatile and
          extractable TCL compound  analyses.

          The  Contractor laboratory must analyze calibration  standards  and
          meet all criteria outlined in Exhibit  E.   After meeting specific
          criteria for both SPCC  and CCC compounds,  a Continuing Calibration
          Data Form must be completed and  submitted.

          Complete all header  information  as  in  part A.   Enter instrument ID,
          date and time of continuing calibration,  the Lab File ID of  the
          continuing  calibration  standard,  and date of initial calibration
          (give inclusive dates  if  initial calibration is performed over more
          than one date).   Using  the appropriate Initial Calibration  (Volatile
          or extractable)  fill in the average relative response factor  (RRF)
          for  each TCL compound.  Report the  relative response factor  from the
          continuing  calibration  standard  analysis.   Calculate the Percent
          Difference  (%D)  for  all compounds.   For  CCC compounds, ensure that
          the  %D is less than  or  equal to  25.0 percent.   After this criterion
          has  been met,  report the  Percent Difference for all TCL and
          surrogate compounds.
                            RRFj  -  RRFC
            % Difference  - 	 x 100
                                RRFj
            where,


            RRFj =  average relative response factor from initial
            r>a~\ i KT^ f~.ion .

                    relative response factor from continuing calibration


                                   B-33                         Rev. 4/89
&WVJ» ~T —   ea- v s^ j.
calibration.

RRFC
•standard.

-------
I.
All continuing calibration standards for extractables (except PCBs) are
analyzed at 80 total ng.  The PCS standards are analyzed at 30 total ng
for mono-heptachlorinated biphenyls, and 60 total ng for octa-
decachlorinated biphenyls.  As on Form VI, if these standards are co-
injected, enter the same Lab File ID in both locations on the form.

2.   Form VII HCA

     The calibration of multicomponent analytes  is verified by a
     comparison of the initial calibration factors to those obtained by
     the periodic analysis of Performance Evaluation Mixtures (PEM).

     Complete the header information as in Part  A.   The  GC Column ID must
     match that on the corresponding Form VI  HCA.   For each PEM injected,
     record the date and time of analysis on  Form VII,  along with amount
     inj ected in. nanograms.   For each of the  three required peaks,  and
     any additional peaks,  enter the retention time of the peak,  in
     minutes and decimal minutes.   For each peak,  calculate the percent
     difference between the  retention time of each peak in the PEM and
     the mean retention time of that peak on  Form VI HCA-2,  according to
     the formula above.  Report this value under "%D" on Form VII HCA.

     Enter initial calibration factor for each peak from Form VI HCA
     under "Initial Cal. Factor".   Enter the  calibration factor from the
     injection listed on Form VII HCA under "Continuing Cal.  Factor."
     Calculate the relative  percent difference according to Ex.  D.  ARO,
     and enter "RPD".

Number_all pages as part A.•

Form VIII

1.   Internal Standard Area  Summary (Form VIII HCV and HCE)

     This form is used to summarize the peak  areas of the internal
     standards added to all  volatile and extractable samples,  blanks,  and
     control matrix spikes.   The data are used to determine when changes
     in internal standard responses will adversely affect quantification
     of target compounds.  This form must be  completed each time a
     continuing calibration is performed, or  when samples are analyzed
     under the same GC/MS tune as an initial  calibration.

     Complete the header information as in part  A.   Enter the Lab File ID
     of the continuing calibration standard,  as  well as  the date and time
     of analysis of the continuing calibration standard.   If samples  are
     analyzed immediately following an initial calibration,  before
     another GC/MS tune and a continuing calibration,  Form VIII shall  be
     completed on the basis  of the internal standard areas of the 50  ug/L
     initial calibration standard for volatiles,  and the 80 ng initial
     calibration standard for extractables.   Use the date-and time  of
     analysis of this standard,  and its Lab File ID and  areas  in place  of
     those of a continuing calibration standard.
                                   B-34
                                                           Rev. 9/88

-------
     From the results of the analysis  of the continuing calibration
     standard,  enter the area measured for each internal standard and its
     retention time under the appropriate column in the row labeled "12
     HOUR STD".   For each internal standard,  calculate the upper limit as
     the area of the particular standard plus 100%  of its area (i.e.,  two
     times the area in the 12 HOUR STD box),  and the lower limit as the
     area of the internal standard minus 50% of its area (i.e.,  one half
     the area in the 12 HOUR STD box).   Report these values in the boxes
     labeled "UPPER LIMIT" and "LOWER  LIMIT" respectively.

     For each sample, blank, and control matrix spike analyzed under a
     given continuing calibration, enter the EPA Sample Number and the
     area measured for each internal standard and its retention time.  If
     the internal standard area is outside the upper or lower limits
     calculated above, flag that area  with an asterisk (*).  The asterisk
     must be placed in the far right hand space of  the box for each
     internal standard area, directly  under the "#" symbol.

     Number all pages as described in  part A.

2.   Analytical Sequence (Form VIII HCA)

     This form is used to record the sequence of analysis of samples and
     standards for all Aroclor/Toxaphene analyses.   Data must be provided
     for the initial calibration standards as well  as the samples in the
     SDG.   If samples in the SDG being reported were analyzed immediately
     after the initial calibration,  list them immediately following the
     calibration standards.   If an older calibration is being used,  list
     the initial calibration standards  on one Form  VIII HCA,  and the
     samples on separate forms,  (as with Form V for volatiles or
     extractables).

     For each Form VIII HCA, complete  all header information as  in part
     A.   Enter the stationary phase of the GC column under "GC Column
     ID".   Begin each Form VIII HCA with calibration standards or
     performance evaluation mixtures associated with the samples.

     For Aroclor/Toxaphene standards,  the following scheme must be used
     to  enter "EPA Sample Number".

                                  EPA Sample  Number
                                     TOXAPH
                                     AR1016
                                     AR1221
                                     AR1232
                                     AR1242
                                     AR1248
                                     AR1254
                                     AR1260
Toxaphene
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
                              B-35
                                                Rev.  9/88

-------
     If Aroclor 1016  and Aroclor 1260 are  combined into one standard
     mixture,  use AR1660 as the "EPA Sample Number".

     Instrument Blanks must be identified  as "IBLK##",  where "##"  may be
     any combination  of numbers or letters needed to  distinguish between
     various instrument blanks in one SDG.

     For each sample,  standard,  control matrix spike,  and blank, enter
     the EPA sample number, lab sample ID, date and time of analysis.

     Every sample, standard, control matrix spike,  method blank, and
     instrument blank must contain the surrogates Tetrachloro-meta-Xylene
     (TMX) and Decachlorobiphenyl (DEC) at the level  specified in  Ex. D.
     ARO.  Calculate  the retention time shift for each of the surrogates
     on both GC columns according to the following formula, and report
     under the appropriate column for %D on Form VIII HCA.
                        RT
RT,
       % Difference  =>
          where:
                           RT
          RT - Mean retention time  from initial  calibration of the
          surrogate
          RTS = Retention time of the  sample  of  interest

     The %D for the surrogates in every sample,  standard,  control matrix
     spike, and blank must be less  than or equal to 0.5%,  as calculated
     above.  Flag each retention time  shift outside the QC limits with an
     asterisk in the last column, under the "*".  If the  surrogates  have
     been diluted out,  and no %D can be calculated,  enter "dil" under
     "%D", and describe in the Case Narrative.

     Number all pages according to  part A.

J.   Form IX

     1.    Extractable GPC Calibration  (Form IX HCE)

          This form reports the results of  the calibration of the gel
          permeation chromatographic apparatus (GPC)  used  to clean-up
          sample extracts.   It reports the  retention times of the four
          GPC calibration compounds, and lists the  phase units associated
          with that calibration.

          Complete  the  header information as  in  Part A.  Enter the
          laboratory's  GPC column identifier  under  "GPC Column ID",  and
          the date  of the GPC calibration.

          For each  of the four GPC  calibration compounds,  enter the
          retention time of that compound during the initial calibration
          under "INITIAL CALIB. RT".   Enter the  retention  time of each
          calibration compound from the calibration verification standard
                              B-36
                            Rev. 4/89

-------
          under "CALIB.  VERIF.  RT".   Calculate the percent difference in
          the two retention times according to the formula for the
          surrogates above,  and enter under "%D".   There are no QC limits
          on percent difference for  Polystrene or  Pentachlorophenol.   The
          QC limit on the other two  GPC calibration compounds is 5%.

          In the lower portion of the form,  list the phase units which
          are associated with this GPC calibration.   Enter the EPA Sample
          No., including the phase unit suffix,  and the Lab Sample ID,
          Lab File ID, and the date  and time analyzed by GC/MS.

          Include a copy of Form IX  HCE for each GPC calibration
          performed that is associated with samples in a data package.
          Number all pages as in Part A.

     2.   Pesticide Retention Times  (Form IX HCA)

          This form reports the retention times and retention time
          windows of the single component pesticide compounds.  While
          these pesticides are not target compounds by the GC/EC methods
          used for Aroclors and Toxaphene,  their retention times are
          reported with the Aroclor/Toxaphene data to ensure that peaks
          for pesticide compounds are not misidentified as Aroclor or
          Toxaphene peaks.

          The retention times of the single compound pesticides must  be
          determined at least once per initial calibration on each
          instrument and each GC column used for sample analysis.  These
          data must be submitted with each group of samples analyzed
          under that initial calibration.

          Retention time windows are determined as a plus/minus
          percentage of the retention time.   For the four BHC compounds
          and Heptachlor, the retention time window is ±1.5% of the
          retention time.  The retention time windows of the remaining 15
          compounds are ±1.0% of the retention time.

          Complete the header information as in Part A.   Enter the GC
          Column ID by stationary phase.   This GC  Column ID must match
          that on Forms II,  VII,  and VIII HCA.  Enter the inclusive dates
          of analysis under "Date(s) of Analysis".

          Enter the retention time of each pesticide under "RT" in
          minutes and decimal minutes,  not minutes and seconds.
          Calculate the retention time windows according to the
          percentages above,  and report as a part  of from/to values,  not
          as a plus/minus percentage.

          No Form X HCA is required  for the analyses of these single
          component pesticides.

          Number all pages as in Part A.

K.   Aroclor/Toxaphene Identification Summary (Form X)

                              B-37                         Rev.  4/89

-------
This form summarizes the tentative and confirmed identity of all TCL
Aroclors/Toxaphene detected in a given sample.  It reports the
retention times of each quantitation peak in the compound on both
columns on which it was analyzed, as well as the retention time
windows of the same peaks in the standard for that compound on both
of these columns.  One copy of Form X is required for each sample,
control matrix spike, or blank in which Aroclors/Toxaphene are
detected.  If none are detected in a given.sample, no copy of Form X
is required for that sample.

Complete the header information as in Part A.  Enter the GC Column
ID (By stationary phase) for each of the two columns, one as GC
Column (1), the other as (2).  Enter the Instrument ID associated
with each GC. column directly below.

For each Aroclor/Toxaphene detected, enter the name of the target
compound on the line under "COMPOUND".  For each of the
quantification peaks selected for that compound (at least 3 peaks
are required), enter the retention time of that peak on each GC
column in the appropriate box under "RT" in decimal minutes.  Enter
the retention time windows of the same peaks in the standard for
that compound.  The retention time window of the standard is defined
as ±1.0% of the mean retention time of the peak determined in the
initial calibration.  The lower value is entered under the "FROM"
column, the upper value under the "TO" column.

Calculate the concentration of each Aroclor or Toxaphene peak
reported on Form X from the calibration factors on each of the two
columns used.  Enter the concentration under the "CONCENTRATION"
column.  Calculate a mean of the 3-5 concentrations for each analyte
on each column that result, and enter this value under the "MEAN
CONCENTRATION" column.  Note:  There will be one mean concentration
per compound in each column.  Report the lower of the two mean
concentrations on Form I HCA.

Calculate the percent difference between the two mean
concentrations, according to the formula below, and enter this value
under the "%D" column.  If the percent difference is greater than
25% of the lower value, flag the result on Form I with one of the
optional flags "X,Y,or Z".  Define your use of the flag in the Case
Narrative.
            % D
   where:

   CT  - lower mean concentration (based on 3-5  peaks)
   CJT = higher mean concentration (based on 3-5 peaks)

If more compounds  are  identified in an  individual sample than can be
reported  on  one copy of  Form X,  then complete  as many additional
copies  of Form X  as necessary,  duplicating all header information,
                          B-38
Rev.

-------
     and numbering the pages as described in Part A.

L.   Diol Cartridge Check (Form XI)

     This form is used to report the results of the Diol  cartridge  check.
     The amount of the spike compound passed through  the  cartridge  and
     removed from the eluent must meet the QC limits  listed on the  form
     before that lot of Diol cartridges may be used for extract clean-up.
     The form also lists all the phase units in the data  package which
     were cleaned using Diol cartridges from this lot.

     Complete all header information as in Part A.  Enter the  lot number
     of the Diol cartridges, and the date of the analysis of the
     cartridge check solution.

     Enter the amount of Aroclor 1254 spiked in the solution in nanograms
     and the amount recovered under  "SPIKE ADDED" and "SPIKE RECOVERY"
     respectively.

     Calculate a percent recovery according to the formula in  Ex. D ARO,
     and enter under "% REC".   This  value must be within  the QC limits of
     80-110%.

     In the lower portion of the form, enter the EPA  Sample No.,
     including phase unit suffix, of each phase unit  which was cleaned
     using a cartridge from this lot.  If there is a  Lab  Sample ID,  enter
     it here as well.  If more than  25 phase units used cartridges  from
     the same lot, duplicate the header information,  and  continue on
     another copy of Form XI,  numbering the pages as  in Part A.
                              B-39
Rev. 4/89

-------
     SECTION IV
DATA REPORTING FORMS
      B-40                         Rev. 9/88

-------
                             1HA
      HIGH CONCENTRATION VOLATILE ANALYSIS DATA SHEET
                                         EPA SAMPLE NO.
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
Phase Type:

Phase weight:
   (g)
Final Extract Volume:

Aliquot Volume:  	
         (mL)
Conversion Factor:
       CAS NO.
      (UL)
         Lab Sample ID:

         Lab File ID:

         Date Received:

         Date Separated:

         Date Extracted:

         Date Analyzed:
   COMPOUND
               CONCENTRATION
                  (mg/Kg)      Q
       74-87-3	Chloromethane	
       74-83-9	Bromome thane	
       75-01-4	Vinyl Chloride	
       75-00-3	Chloroethane	._
       75-09-2	Methylene Chloride	.
       67-64-1	Acetone	'
       75-15-0	Carbon Disulfide   	
       75-35-4	1,1-Dichloroethene	
       75-34-3	1,1-Dichloroethane	
       540-59-0	1,2-Dichloroethene (total)
       67-66-3	Chloroform	'
       107-06-2—	1,2-Dichloroethane	
       78-93-3	2-Butanone	
       71-55-6	1, l, 1-Trichloroethane	
       56-23-5	Carbon Tetrachloride	
       108-05-4	Vinyl Acetate	
       75-27-4	Bromodichloromethane	
       78-87-5	1,2-Dichloropropane	
       10061-01-5	cis-l,3-Dichloropropene	
       79-01-6	Trichloroethene	
       124-48-1	Dibromochloromethane	
       79-00-5	1,1,2-Trichloroethane	
       71-43-2	Benzene	
       10061-02-6	trans-l,3-Dichloropropene_
       75-25-2	Bromoform	~
       108-10-1	4-Methyl-2-Pentanone	
       591-78-6	2-Hexanone	^_
       127-18-4	Tetrachloroethene	
       79-34-5	l,l,2,2-Tetrachloroethane_
       108-88-3	Toluene	~
       108-90-7	Chlorobenzene	
       100-41-4—	Ethylbenzene	
       100-42-5	Styrene	
       1330-20-7	Xylene (total)	
                                FORM I HCV
                                                 Rev.  9/8;

-------
                             1KB
    •HIGH CONCENTRATION EXTRACTABLE ANALYSIS DATA SHEET
                                         EPA SAMPLE NO.
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.':
SDG No.:
Phase Type:

Phase weight:
   (g)
Final Extract Volume:

Injection Volume:  	

Conversion Factor: 	

pH:  	


       CAS NO.
     (UL)
         Lab Sample ID:

         Lab File ID:

         Date Received:

         Date Separated:

         Date Extracted:

         Date Analyzed:
   COMPOUND
               CONCENTRATION
                   (mg/Kg)
       108-95-2	Phenol	
       111-44-4	bis (2-Chloroethyl) ether	
       95-57-8	2-Chlorophenol	
       541-73-1-	1,3-Dichlorobenzene	
       106-46-7	1,4-Dichlorobenzene	
       100-51-6	-Benzyl alcohol       	
       95-50-1	1,2-Dichlorobenzene	
       95-48-7	2-Methylphenol__	
       108-60-1--	bis(2-Chlorbisopropyl)ether_
       106-44-5	4-Methylphenol	
       621-64-7	N-Nitroso-di-n-propylamine	
       67-72-1	Hexachloroethane	
       98-95-3	Nitrobenzene	
       78-59-1	Isophorone	
       88-75-5	2-Nitrophenol	
       105-67-9	2, 4-Dimethylphenol	
       65-85-0	Benzoic acid	
       111-91-1	bis(2~Chloroethoxy)methane	
       120-83-2	2,4-Dichlorophenol	
       120-82-1	1,2,4-Trichlorobenzene	
       91-20-3—	Naphthalene	
       106-47-8	4-Chloroaniline	
       87-68-3	Hexachlorobutadiene	
       59-50-7	4-Chloro-3-methylphenol	
       91-57-6	2-Methylnaphthalene	
       77-47-4	Hexachlorocyclopentadiene
       88-06-2	2, 4, 6-Trichlorophenol	
       95-95-4	2,4,5-Trichlorophenol	
       91-58-7	2-Chloronaphthalene	
       88-74-4	2-Nitroaniline	
       131-11-3	Dimethylphthalate	
       208-96-8	Acenaphthylene	
       606-20-2	2 , 6-Dinitrotoluene	
                                FORM I HCE-1
                                                Rev. 9/88

-------
                             1HC
     HIGH CONCENTRATION EXTRACTABLE ANALYSIS DATA SHEET
                                                             EPA  SAMPLE  NO.
Lab Name:

Lab Code:
                    Case No. :
Contract:

 SAS No.:
SDG No.:
Phase Type:

Phase weight:
                       (g)
Final Extract Volume:

Injection Volume:

Conversion Factor:

pH:  	


       CAS NO.
                             (mL)
                         (uL)
         Lab Sample 10:

         Lab File ID:

         Date Received:

         Date Separated:

         Date Extracted:

         Date Analyzed:
                       COMPOUND
               CONCENTRATION
                  (mg/Kg)      Q
       99-09-2	3-Nitroaniline	
       83-32-9	—Acenaphthene	
       51-28-5	2,4-Dinitrophenol	
       100-02-7	4-Nitrophenol_	
       132-64-9	Dibenzofuran	
       121-14-2-	2 ,4-Dinitrotoluene	
       84-66-2	Diethylphthalate	
       7005-72-3	4-Chlorophenyl-phenylether_
       86-73-7	Fluorene   	
       100-01-6	4-Nitroaniline	
       534-52-1	4, 6-Dinitro-2-methylphenol_
       86-30-6	N-Nitrosodiphenylamine (1)~
       101-55-3	4-Bromophenyl-phenylether_^
       319-84-6	alpha-BHC	
       118-74-1	Hexachlorobenzene	
       319-85-7	beta-BHC	
       87-86-5	Pentachlorophenol	
       58-89-9	gamma-BHC (Lindane)	
       85-01-8	Phenanthrene	
       120-12-7	Anthracene	
       319-86-8	delta-BHC	
       76-44-8	Heptachlor	
       309-00-2	Aldrin	
       84-74-2	
         5-44-0	            	
           -57-3	Heptachlor epoxide_
           5-18-8	Monochlorobiphenyl	

             1-8	Trichlorobiphenyl
             2-9	Tetrachlorobiphenyl_
             -0	Pyrene	'_
       5103-74-2	gamma-Chlordane	
     (1)  - Cannot be separated from Diphenylamine

                                FORM I HCE-2
                                                                     Rev. 9/81

-------
                             1HD
     HIGH CONCENTRATION EXTRACTABLE ANALYSIS DATA SHEET
                                         EPA SAMPLE NO.
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
Phase Type:

Phase weight:
   (g)
Final Extract Volume:

Injection Volume:  	

Conversion Factor: 	

pH:  	


       CAS NO.
         (mL)
     (UL)
         Lab Sample ID:

         Lab File ID:

         Date Received:

         Date Separated:

         Date Extracted:

         Date Analyzed:
   COMPOUND
               CONCENTRATION
                   (mg/Kg)
       959-98-8	Endosulfan I	
       5103-71-9	alpha-Chlordane	
       25429-29-2	Pentachlorobiphenyl_

       60-57-1	Dieldrin

       72-20-8	      	

       72-54-8	.4,4'-T
       28655-71-2	•—!
       85-68-7	
       1031-07-8	Endosulfan sulfate
       50-29-3	4 , 4 ' -DDT

       56-55-3	                  	
       72-43-5	Methoxychlor     	
       218-01-9	Chrysene	
       55722-26-4	Octachlorobiphenyl	
       91-94-1	3,3 ' -Dichlorobenzidine	
       117-81-7	bis(2-Ethylhexyl)phthalate
       53742-07-7	
       2051-24-3	
       117-84-0	Di-n-
                            i(b) fluoranthene	
                             .k)fluoranthene	
       50-32-8	Benzo (a) pyrene	
       193-39-5	Indeno (1,2, 3-cd) pyrem
       53-70-3	Dibenz (a,h) anthracene^
       191-24-2	Benzo (g,h,i)perylene
                                FORM  I HCE-3
                                                Rev.  9/88

-------
                             1HE
      HIGH CONCENTRATION VOLATILE ANALYSIS DATA SHEET
                TENTATIVELY IDENTIFIED COMPOUNDS
                                      EPA SAMPLE NO.
Lab Name:
Lab Code:
Phase Type:
Contract: |
Case No.: SAS No.: SDG No.:
_ 	 _. . Lab Sample TD:
Phase weight:
(9)
Final Extract Volume:
Aliquot Volume:  	
      (mL)
   (uL)
Conversion Factor:
Number TICs found:
Lab File ID:
Date Received:
Date Separated:
Date Analyzed:
CAS NUMBER
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14. .
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.

COMPOUND NAME





,

























RT































EST. CONC.
(mg/Kg)































Q



1
1


|
1




















1
	 |
                                FORM I HCV-TIC
                                             Rev."  9/8i

-------
                             1HF
   HIGH CONCENTRATION EXTRACTABLE ANALYSIS DATA SHEET
                TENTATIVELY IDENTIFIED COMPOUNDS
                                         EPA SAMPLE NO
Lab Name:
Lab Code:
Case No.:
Contract:
 SAS No.:
SDG No.:
Phase Type:
Phase weight:
   (g)
Final Extract Volume:
Injection Volume:
Conversion Factor:
pH:  	
Number TICs found:
         (mL)
     (uL)
         Lab Sample ID:
         Lab File ID:
         Date Received:
         Date Separated:
         Date Extracted:
         Date Analyzed:
CAS NUMBER
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
•23.
24.
25.
26.
27.
28.
29.
30.

COMPOUND NAME































RT































EST. CONG.
(mg/Kg)






• .
























Q































                                 FORM  I .HCE-TIC
                                                  Rev.  9/88

-------
                             1HG
     HIGH CONCENTRATION AROCLOR ANALYSIS DATA SHEET
                                      EPA SAMPLE NO.
Lab Name:
Lab Code:

Case No. :
Contract :
SAS No. : S
                                                        SDG No.:
Phase Type:

Phase weight:
(g)
Final Extract Volume:

Injection Volume:

Conversion Factor:

pH:  	


       CAS NO.
      (mL)
  (uL)
COMPOUND
       8001-35-2	
       12 674-11-2	Aroclor-]
       11104-28-2	Aroclor-:
       11141-16-5	1
       53469-21-9	;
       12672-29-6	
       11097-69-1	
       11096-82-5	Aroclor-1260"
Lab Sample ID:

Date Received:

Date Separated:

Date Extracted:

Date Analyzed:
                       Sulfur Clean-up:
      CONCENTRATION
         (mg/Kg)
                                FORM I HCA
                                           Rev.  9/88

-------
                              2HA
          HIGH CONCENTRATION VOLATILE SURROGATE RECOVERY
Lab Name:

Lab Code:
Case No.:
Contract;

 SAS No.:
SDG No.
| EPA
j SAMPLE NO.
oil
02|
03|
04|
05|
061
07|
08|
09|
10 |
HI
121

141
151
16 |
17 |
18]
191
20|
211
22J
23|
24|
25|
26|
27|
28|
291
30|
SI
(TOL) #






























S2
(BFB)#

















•












S3
(DCE)#






























OTHER
'






























TOT
OUT






























                                                QC LIMITS
               SI  (TOL) = Toluene-d8              (50-160)
               S2  (BFB) = Bromofluorobenzene      (50-160)
               S3  (DCE) = l/2-Dichloroethane-d4   (50-160)

               # Column to be used to flag recovery values

               D Surrogates diluted out

     Method blanks are required to meet the QC limits.   For samples
     and control matrix spikes, the QC limits are advisory.
page 	 of 	
                                FORM II HCV
                                              Rev. 9/88

-------
                               2HB
       HIGH CONCENTRATION  EXTRACTABLE SURROGATE RECOVERY
Lab Name:

Lab Code:
Case No.:
Contract;

 SAS No.:
SDG No,
       01!
       02|
       03
       04 |
       05 I
       06)
       07)
       08
       09
       10|'
       11
       12 |
       13 |
       14 |
       15|
       16
       17 I
       18 '
       191.
       20
       21|
       22 |
       23
       24;
       25|
       26
       27|
       28
       29
EPA
SAMPLE NO.






























SI
(NBZ)#






























S2
(FBP) #






























S3
(TPH) #






























S4
(PHL) *






























S5
(2FP)#






























S6
(TBP) f























"






OTHER






























TOT
OUT






























                SI  (NBZ) = Nitrobenzene-dS
                S2  (FBP) = 2-Fluorobiphenyl
                S3  (TPH) - Terphenyl-dl4
                S4  (PHL) » Phenol-d5
                S5  (2FP) = 2-Fluorophenol
                S6  (TBP) = 2,4,6-Tribromophenol
                             QC LIMITS
                               (20-140)
                               (20-140)
                               (20-150)
                               (20-140)
                               (20-140)
                               (10-140)
                # Column to be used to  flag  recovery values
                D Surrogates diluted  out
     Method blanks are required to meet the  QC  limits.   For samples
     and control matrix spikes, the QC  limits are  advisory
page 	 of 	
                                FORM  II HCE
                                              Rev. 9/88

-------
                                                                              •\
                              2HC
       HIGH CONCENTRATION AROCLOR SURROGATE RECOVERY
Lab Name:

Lab Code:
GC Column ID (1):
Case No.:
Contract:

 SAS No.:
SDG No.
             GC Column ID (2):
1
t EPA
| SAMPLE NO.
1 	
Oil
021
031
04|
051
06 |
07 |
08 |
091
101
111
121
131
14|
15|
161
171
18|
191
20|
211
22|
231
24|
25|
26|
27|
28|
29|
30|
COL.l |
SI |
(TMX)#!





























COL. 2
SI
(TMX)#





























COL.l
S2
(DEC) f






























COL. 2
S2
(DEC) f
'





























                                                    QC  LIMITS
                SI  (TMX) = Tetrachloro-meta-Xylene    (40-120)
                S2  (DEC) - Decachlorobiphenyl         (40-120)

                # Column to be used to  flag  recovery values

                D Surrogate diluted out
page 	 of 	
                                 FORM II  HCA
                                                Rev.  9/88

-------
                             3 HA
    HIGH CONCENTRATION VOLATILE CONTROL MATRIX SPIKE RECOVERY
Lab Name:
Lab Code:
Case No.:
Contract:
 SAS No. :
SDG No.
COMPOUND
____ __
	 , , 	
1 , 1-Dichloroethene
Trichloroethene
Benzene
Toluene ,
Chlorobenzene

SPIKE
ADDED
(ntg/Kg)






CMS
CONCENTRATION
(rag/Kg)






CMS
%
REC |






QC. |
LIMITS |
REC. |
60-150|
60-150|
60-1501
60-1501
60-150|
1
       # Column to be used to flag recovery values with an asterisk
       * Values outside of QC limits
       Spike Recovery:	 out of     '   outside limits
COMMENTS:
                                FORM III HCV
                                               Rev. 9/88

-------
                             3HB
    HIGH CONCENTRATION EXTRACTABLE CONTROL MATRIX SPIKE RECOVERY
Lab Name:
Lab Code:
Case No.:
Contract:
 SAS No.:
SDG No.
COMPOUND
Phenol
2 -Chlorophenol
1, 4-Dichlorobenzene
N-Nitroso-di-n-prop. (1)
1,2, 4-Trichlorobenzene
4-Chloro-3-methylphenol
Acenaphthene
4-Nitrophenol
2 , 4-Dinitrotoluene
Pent achl or opheno 1
Heptachlor
Pyrene
Dieldrin

SPIKE
ADDED
(mg/Kg)














CMS
CONCENTRATION
(mg/Kg)














CMS | QC. |
% | LIMITS |
REC #| REG. |
110-1201
I 10-120 j
130-1401
|30-140|
130-1401
1 10-1201
| 30-140|
110-1201
| 30-140|
|10-120|
| 30-140)
130-1401
130-1401
1 1
        (1) N-Nitroso-di-n-propylamine
       # Column to be used to flag recovery values with an asterisk
       * Values outside of QC limits
       Spike Recovery:	 out of 	 outside limits
COMMENTS:
                                FORM III HCE
                                             Rev. 9/88

-------
                             3HC
    HIGH CONCENTRATION AROCLOR CONTROL MATRIX SPIKE RECOVERY
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No,


COMPOUND
Aroclor 1254

| SPIKE
| ADDED
1 (ng/Kg)
1
1
| CMS
| CONCENTRATION
| (mg/Kg)
1
i
CMS
%
REC


1 QC.
| LIMITS
| REC.
j none
1
COMMENTS:
                                FORM III HCA
                                               Rev. 9/88

-------
                              4HA
       HIGH CONCENTRATION VOLATILE METHOD BLANK SUMMARY
Lab Name:
Lab Code:
Lab Sample ID:
Lab File ID:
Instrument ID:
Case No.:
Contract:
 SAS No.:
SDG No.:
                          Date Analyzed:
                          Time Analyzed:
      THIS METHOD BLANK APPLIES TO THE FOLLOWING SAMPLES AND  CMS:
[ EPA
| SAMPLE NO.
j
on
021
031
041
051
06|
07 t
081
09 (
10 [
HI
121
13|
141
151
161
171
18 |
191
20|
211
221
23|
241
25|
261
27|
281
291
30|
LAB
SAMPLE ID










.



















LAB
FILE ID
'





























TIME
ANALYZED






























COMMENTS:
page 	 of 	
                                FORM  IV HCV
                                               Rev. 9/88

-------
                                4KB
       HIGH  CONCENTRATION EXTRACTABLE METHOD BLANK SUMMARY
 Lab Name:
 Lab Code:
, Lab Sample  ID:
 Lab File  ID:
„ Instrument  ID:
Case No.:
Contract:
 SAS No.:
SDG No.:
                          Date Extracted:
                          Date Analyzed:
                          Time Analyzed:
         THIS METHOD  BLANK APPLIES  TO  THE  FOLLOWING SAMPLES AND CMS:
| EPA
| SAMPLE NO.
on
02j
031
04|
051
061
07]
081
091
101
HI
12|
131
141
15|
161
17|
181
191
20|
211
221
231
24!
251
26|
27|
28|
291
30|
LAB
SAMPLE ID























.






LAB
FILE ID











'


















DATE
ANALYZED






























TIME
ANALYZED






























  COMMENTS:
 page
         of
                                  FORM TV  HCE
                                               Rev. 9/88

-------
                               4HC
      HIGH CONCENTRATION AROCLOR METHOD BLANK SUMMARY
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.
Lab Sample ID:
Date Analyzed  (1):

Time Analyzed  (1):

Instrument ID  (1):

GC Column ID   (1):

Sulfur Clean-up:
                          Date  Extracted:

                          Date  Analyzed (2)

                          Time  Analyzed (2)

                          Instrument ID (2)

                          GC  Column ID  (2)
         THIS METHOD BLANK APPLIES  TO THE FOLLOWING SAMPLES AND CMS:
01|.
02 |
03
04
05 |
06|
o?i;
08
091
10 |
"I.
12;

14
15|
"I.

is i;
19|.
20|
21|
22
23|.
24|.
25|.
26j
27|
28
29
30|
| EPA
| SAMPLE NO.
1
1
1
1
1
1 '
1
!
1
1
1
1
I
1
1
I
|
I
1
1
1
I
1
1
1
1
1
I
1
1
LAB
SAMPLE ID







, '
"





















DATE
ANALYZED 1






.























TIME
ANALYZED 1


















'











DATE
ANALYZED 2






























TIME
ANALYZED . 2






























COMMENTS:
page 	 of 	
                                FORM IV HCA-1
                                                 Rev.  9/88

-------
                              4HD
      HIGH CONCENTRATION AROCLOR INSTRUMENT BLANK SUMMARY
Lab Name:
Lab Code:

Case No. :
Contract :
SAS No. : S
Lab Sample ID:
Date Analyzed  (1):
Time Analyzed  (1) :
Instrument ID  (1):
GC Column ID   (1) :
          SDG No.:
Date Prepared:
Date Analyzed (2):
Time Analyzed (2):
Instrument ID (2):
GC Column ID  (2):
       THIS INSTRUMENT BLANK APPLIES TO THE FOLLOWING SAMPLES  AND CMS:
| EPA
I SAMPLE NO.
Oil
021
; 031
: 041
05!
06|
07|
, 081
091
10!
HI
121
131
141
151
* 16|
17!
181
191
201
211
221
23|
241
> 251
261
27|
28|
291
301
LAB
SAMPLE ID






























DATE
ANALYZED 1






























TIME
ANALYZED 1






























DATE
ANALYZED 2






























TIME
ANALYZED 2



























«


COMMENTS:
page 	 of 	
                               FORM IV HCA-2
                      Rev. 9/88

-------
                              5HA
      HIGH CONCENTRATION VOLATILE GC/MS TUNING AND MASS
            CALIBRATION - BROMOFLUOROBENZENE (BFB)
Lab Name:
Lab Code:
Case No.:
Contract:
 SAS No.:
SDG No.:
Lab File ID:
Instrument ID:
                       BFB Injection Date:
                       BFB Injection Time:
1
1 ra/e
i
| 50
I 75
1 95
1 96
| 173
1 174
| 175
| 176
| 177
1
ION ABUNDANCE CRITERIA
15.0 - 40.0% Of mass 95
30.0 - 60.0% of mass 95
Base peak, 100% relative abundance
5.0 - 9.0% of mass 95
Less than 2.0% of mass 174
Greater than 50.0% of mass 95
5.0 - 9.0% of mass 174
Greater than 95.0%, but less than 101.0% of mass 174
5.0 - 9.0% of mass 176

% RELATIVE
ABUNDANCE




( )1

( )1
( )1
( )2

        1-Value is % mass 174
                          2-Value is % mass 176
THIS TUNE APPLIES TO THE FOLLOWING SAMPLES, CMS, BLANKS, AND STANDARDS: .
| EPA
| SAMPLE NO.
i — -
1
on
021
03|
04 |
051
061
071
081
09|
10|
HI
121
13|
14|
151
161
17!
181
19|
20|
211
221
LAB
SAMPLE ID






















LAB
FILE ID






















DATE
ANALYZED






















TIME
ANALYZED






















page 	 of 	
                                FORM V HCV
                                                Rev.  9/88

-------
                               5HB
        HIGH CONCENTRATION EXTRACTABLE GC/MS TUNING AND MASS
         CALIBRATION  -  DECAFLUOROTRIPHENYLPHOSPHINE (DFTPP)
Lab Name:
                 Contract:
Lab Code:
Case No.:
SAS No.:
SDG No.
Lab File ID:

Instrument ID:
                     DFTPP Injection Date:

                     DFTPP Injection Time:
1
1 ^^ »
I 51
| 68
| 69
| 70
1 127
| 197
] 198
| 199
| 275
( 365
| 441
| 442
1 443
i
ION ABUNDANCE CRITERIA
30.0 - 60.0% of mass 198
Less than 2.0% of mass 69
Mass 69 relative abundance
Less than 2.0% of mass 69
40.0 - 60.0% of mass 198
Less than 1.0% of mass 198
Base Peak, 100% relative abundance
5.0 to 9.0% of mass 198
10.0 - 30.0% of mass 198
Greater than 1.00% of mass 198
Present, but less than mass 443
Greater than 40.0% of mass 198
17.0 - 23.0% of mass 442

% RELATIVE
ABUNDANCE

( ) 1

( ) 1








r )2

         1-Value  is  % mass  69
                          2-Value is % mass 442
THIS TUNE APPLIES TO THE  FOLLOWING  SAMPLES,  CMS,  BLANKS,  AND STANDARDS:
| EPA
j SAMPLE NO.
on
02|
03|
04|
05 |
061
07|
081
09 |
10 |
HI
12|
131
141
15 |
16 |
171
18 |
191
20!
211
22 |
LAB
SAMPLE ID
t





















LAB
FILE ID






















DATE
ANALYZED






















• TIME
ANALYZED














:,







page 	 of 	
                                FORM V HCE
                                                Rev.  9/88

-------
                              6HA
     HIGH CONCENTRATION VOLATILE INITIAL CALIBRATION DATA
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
Instrument ID:
       Calibration Date(s):
Min RRF for SPCC(#) = 0.300  (0.250 for Bromoform) Max %RSD  for  CCC(*)  = 30.0%
ILAB FILE ID: RRF20
RRF50 =
|RRF100= RRF150= RRF200=
I
1 1
| COMPOUND |RRF20
i " I
| Chloromethane #
i Bromomethane I
i Vinyl Chloride * .
1 Chloroethane j
(Methylene Chloride |
I Acetone I
j Carbon Disulfide
1 1 , 1-D ichl or oethene *
1 1 , 1-Dichloroethane #
1 1,2-Dichloroethene (total) |
I Chloroform *
1 1 , 2-Dichloroethane
(2-Butanone
11,1, 1-Trichloroethane
[Carbon Tetrachloride
[Vinyl Acetate
j Bromodichloromethane
j 1 , 2 -Dichloropropane *
I cis-l , 3-Dichloropropene
1 Trichloroethene
j Dibromochloromethane
| 1, 1, 2-Trichloroethane
I Benzene
j trans-1, 3-Dichloropropene
| Bromoform •
| 4-Methyl-2-Pentanone
| 2-Hexanone






it






i


! Tetrachloroethene
| 1,1,2, 2-Tetrachloroethane #
j Toluene . *
| Chlorobenzene #
| Ethylbenzene *
| Styrene |
IXylene (total) !
| Toluene-d8
| Bromof luorobenzene
j 1 , 2-Dichloroethane-d4
j
RRF50






































RRF100














,
%






















RRF150






































RRF200





































RRF







































% 1
RSD |
i
#
I
*
1
1
i
I
*
. #
i
*
1
1
=4
1
*
	 1
1
I
1
1
1
§
1
1
	 1
i
*
#
*
I
i
1
1
1
1
1
                                 FORM VI  HCV
                                                  Rev.  9/

-------
                                6HB
       HIGH CONCENTRATION EXTRACTABLE INITIAL CALIBRATION DATA
  Lab  Name:

  Lab  Code:
             Case No.:
Contract:

 SAS No.:
SDG No.:
  Instrument ID:
                    Calibration Date(s):
l  Kin RRF for SPCC(#)  = 0.050
                                           Max %RSD for CCC(*) = 30.0%
| LAB FILE ID:
                        RRF50 =
                        RRF160="
        RRF80
             'I
1
| COMPOUND
1 Phenol *
1 bis (2-Chloroethyl) ether
| 2-Chlorophenol
I 1 , 3-Dichlorobenzene
|l,4-Dichlorobenzene '
I Benzyl alcohol
| 1,2-Dichlorobenzene
I 2-Methylphenol
j bis(2-Chloroisopropyl> ether
! 4-Methylphenol
|N-Nitroso-di-n-propylamine j
I Hexachloroethane
j Nitrobenzene
I Isophorone
1 2-Nitrophenol --i
1 2 , 4-Dimethylphenol
[Benzoic acid
| bis (2-Chloroethoxy) methane
j 2 , 4-Dichlorophenol •>
| 1 , 2 , 4-Trichlorobenzene
I Naphthalene
j 4-Chloroaniline
(Hexachlorobutadiene '
i 4-Chloro-3-methylphenol *
I 2-Methylnaphthalene
JHexachlorocyclopentadiene :
I 2 , 4 , 6-Trichlorophenol *
j 2 , 4 , 5-Trichlorophenol
| 2-Chloronaphthalene
|2-Nitroaniline .
I Dimethylphthalate
.| Acenaphthylene
j 2 , 6-Dinitrotoluene
j 3-Nitroaniline
I Acenaphthene *
I 2 1 4-Dinitrophenol $
|4-Nitrophenol i
1 ' 1

RRF50
t




-------
                              6HC
      HIGH CONCENTRATION EXTRACTABLE INITIAL CALIBRATION DATA
Lab Name:
                 Contract::
Lab Code:
Case No.
SAS No.:
SDG No.:
Instrument ID:
       Calibration Date(s):
Min RRF for SPCC(#) - 0.050
                              Max %RSD for CCC(*) - 30.0%
        [LAB FILE ID:  RRF50
                         RRF80
I RRF160=
1
1
| COMPOUND |
	 - a • -
| Dibenzofuran
| 2 , 4-Dinitrotoluene
I Diethylphthalate
| 4-Chlorophenyl-phenylether
j Fluorene
I 4-Nitroaniline
| 4 , 6-Dinitro-2-methylphenol
| N-Nitrosodiphenylaraine (1) *
| 4-Bromophenyl-phenylether
1 alpha-BHC
| Hexachlorobenzene
| beta-BHC
1 Pentachlorophenol
j gamma-BHC ( Lindane }
| Phenanthrene
I Anthracene
I delta-BHC
I Heptachlor
I Aldrin
j Di-n-butylphthalate
| Fluoranthene
| Heptachlor epoxide
| Pyrene
| gamma-Chlordane
| Endosulfan I
| alpha-Chlordane
|4,4 '-DDE
I Dieldrin
1 Endrin
| Endosulfan II
|4,4'-DDD
| Butylbenzylphthalate
I Endosulfan sulfate
|4,4 '-DDT
j Endrin ketone
| Benzo (a) anthracene
| Methoxychlor
I


RRF50







*




k







k



















RRF80 |








































RRF160








































RRF








































%
RSD






































 (1) Cannot  be separated from Diphenylamine

                                 FORM VI HCE-2
                                                  Rev.9/88

-------
                              6HD
      HIGH CONCENTRATION EXTRACTABLE INITIAL CALIBRATION DATA
Lab Name:

Lab Code:
                    Case No . :
Contract:

 SAS No.:
                                                        SDG No.:
Instrument ID:
                           Calibration Date(s) :
Min RRF for SPCC(#) - 0.050
                                                  Max %RSD for CCC(*} = 30.C
ILAB FILE ID: RRFSO -
| RRF160-
1
1
| COMPOUND
1 Chrysene
13,3' -Dichlorobenzidine
| bis (2-Ethylhexyl) phthalate
j Di-n-octylphthalate '
| Benzo (b) f luoranthene
| Benzo (k) f luoranthene
| Benzo (a) pyrene '
I Indeno (1,2, 3-cd) pyrene
j Dibenz (a,h) anthracene
I Benzo (g,h,i)perylene

[ Nitrobenzene-d5
} 2 -Fluor ob ipheny 1
| Terphenyl-dl4
1 Phenol-d5
I 2-Fluorophenol
1 2 , 4 , 6-Tribromophenol

ILAB FILE ID: RRFIO =
I RRFSO =
1
1
1
I Monochlorobiphenyl
j Dichlorobiphenyl
I Trichlorobiphenyl
I Tetrachlorobiphenyl
j Pentachlorobiphenyl
j Hexachlorobiphenyl
j Heptachlorobiphenyl

ILAB FILE ID: RRF20 =
1 RRF100=
1
1
I
I Octachlorob ipheny 1
j Nonachlorobiphenyl
j Decachlorobiphenyl
1



| RRFSO



k


k















RRFIO











RRF20



RRF 8l


RRFSO


















RRF3(



RRF30








RRF6I


RRF60



3 =


IRRF160


















) =



RRFSO








3 -


RRF100






RRF






















RRF











RRF






RSD



4


4














%
RSD











%
RSD



                                FORM VI HCE-3
                                                                    Rev.9/8£

-------
                              6HE
  HIGH CONCENTRATION INITIAL CALIBRATION OF MULTICOMPONENT ANALYTES
Lab Name:_

Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
Instrument ID:
                  GC Column ID:
 * Denotes required peaks

page 	 of 	
COMPOUND
Toxaphene
Date :

Time:

Aroclor 1016
Date:

Time:

Aroclor 1221
Date:

Time:

Aroc.lor 1232
Date:

Time:

Aroclor 1242
Date :

Time:

Aroclor 1248
Date:

Time:

Aroclor 1254
Date:

Time:

Aroclor 1260
Date:

Time:


AMOUNT
(ng)









































PEAK
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5

RT









































CALIBRATION
FACTOR
	








































                                FORM VI HCA-1
                                                Rev. 9/88

-------
                              6HF
  HIGH CONCENTRATION INITIAL CALIBRATION OF MULTICOMPONENT ANALYTES
Lab Name:,
Lab Code:
Case No.:
Contract:.
 SAS No.:
SDG No.:
Instrument ID:
                  GC Column ID:
Date(s) Analyzed:
COMPOUND
Toxaphene




Arolcor 1016




Aroclor 1221




Aroclor 1232




Aroclor 1242




Aroclor 1248




Aroclor 1254




Aroclor 1260





PEAK
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5

MEAN
RT









































MEAN CAL.
FACTOR









































%
RSD









































r
(1)









































INTCP
(1)









































r
(2)









































INTCP
(2)









































 * Denotes required peaks
page 	 of 	
                                FORM VI HCA-2
                                                 Rev. 4/

-------
                              7HA
   HIGH CONCENTRATION VOLATILE CONTINUING CALIBRATION CHECK
Lab Name:
Lab Code:
Case No.:
Contract:
 SAS No.:
Instrument ID:
Lab File ID:
        Calibration Date:
SDG No.:
                    Time:
        Init. Calib. Date(s):
Min RRF50 for SPCC{#) - 0.300  (0.250 for Bromoform) Max  %D  for CCC(*)  - 25.0%
COMPOUND
Chloromethane •
Bromomethane
Vinyl Chloride *
Chloroethane
Methylene Chloride
Acetone
Carbon Disulfide
1,1-Dichloroethene *
1,1-Dichloroethane i
1,2-Dichloroethene (total)
Chloroform *
1, 2-Dichloroethane
2-Butanone
1 , 1 , 1-Trichloroethane
Carbon Tetrachloride
I Vinyl Acetate
Bromodichloromethane
1, 2-Dichloropropane '
cis-l, 3-Dichloropropene
Trichloroethene
Dibromochlorome thane
1,1, 2-Tr ichloroethane
Benzene
trans-1 , 3-Dichloropropene
Bromoform 1
4 -Methy 1-2 -Pentanone
2-Hexanone
Tetrachloroethene
1,1,2,2-Tetrachloroethane j
Toluene ^
Chlorobenzene ;
| Ethylbenzene *
Styrene
Xylene (total)

Toluene-d8
Bromofluorobenzene
1 , 2 -Dichloroethane-d4

RRF
f

it
.



IT
s

>f .






k






'•



l-
k
5
ir







RRF50







































|
%D |
,, 	 j
i
!
4
\
\


i
1

4




*

4






«



\
i
i
4
• 1
I




1
                                 FORM VII  HCV
                                                 Rev.  9/88

-------
                              7HB
     HIGH CONCENTRATION EXTRACTABLE CONTINUING CALIBRATION CHECK
Lab Name:
Lab Code :
Instrument ID:

Case No. :
Contract:
SAS No. :
Calibration Date:

SDG No . :
Time:
Lab File ID:
Init. Calib. Date(s):
Min RRF80 for SPCC(#) = 0.050
                        Max %D for CCC(*) = 25.0%
1
| COMPOUND
1 Phenol
j bis (2-Chloroethyl) ether
| 2-Chlorophenol
j 1, 3-Dichlorobenzene
j 1,4-Dichlorobenzene
[Benzyl alcohol
j 1, 2-Dichlorobenzene
| 2-Methylphenol
| bis (2-Chloroisopropyl) ether
| 4-Methylphenol
j N-Nitroso-di-n-propylamine
| Hexachloroe thane
j Nitrobenzene
| Isophorone
I 2-Nitrophenol *
j 2 , 4-Dimethylphenol
JBenzoic acid
| bis (2-Chloroethoxy) methane
j 2 , 4-Dichlorophenol *
j 1 , 2 , 4-Trichlorobenzene
| Naphthalene
| 4-Chloroaniline
j Hexachlorobutadiene *
| 4-Chloro-3-methylphenol *
\ 2-Methylnaphthalene
j Hexachlorocyclopentadiene \
j 2 , 4 , 6-Trichlorophenol <
(2,4, 5-Trichlorophenol
| 2-Chloronaphthalene
j 2-Nitroaniline
| Dimethylphthalate
1 Acenaphthylene
| 2 , 6-Dinitrotoluene
j 3-Nitroaniline
| Acenaphthene *
| 2 , 4 -Dinitrophenol 1
1 4-Nitrophenol 1
1
RRF
i



,

















it
t

1
t







^
'•
''•

RRF80






































%D














*



*



*
*
1
1
*







*
1
#
1
                                FORM VII HCE-1
                                        Rev.  1/89

-------
                              7HC
   HIGH CONCENTRATION EXTRACTABLE CONTINUING CALIBRATION CHECK
Lab Name:
                                     Contract:
Lab Code:
                    Case No.:
                                      SAS No.:
Instrument ID:
Lab File ID:
                            Calibration Date:
SDG No.:
 Time:
                            Init. Calib. Date(s):
Min RRF80 for SPCC(#) = 0.050
                                                    Max %D for CCC(*) = 25.0%
1
| COMPOUND
I Dibenzofuran
1 2 , 4-Dinitrotoluene
1 Diethylphthalate
| 4-Chlorophenyl-phenylether
I Fluorene
j 4-Nitroaniline
j 4 , 6-Dinitro-2-methylphenol
j N-Nitrosodiphenylamine (1) <
| 4-Bromophenyl-phenylether
I alpha-BHC
| Hexachlorobenzene
| beta-BHC
! Pentachlorophenol *
j ganona-BHC ( Lindane )
j Phenanthrene
j Anthracene
1 delta-BHC
j Heptachlor
lAldrin
| Di-n-butylphthalate
| Fluoranthene •>
j Heptachlor epoxide
| Pyrene
| gamma-Chlordane
| Endosulfan I
| alpha-Chlordane
14,4' -DDE
1 Dieldrin
| Endrin
| Endosulfan II
14,4' -ODD
| Butylbenzylphthalate
| Endosulfan sulfate
1 4, 4 '-DDT
j Endrin ketone
| Benzo(a) anthracene
I Methoxychlor
1

RRF |







































RRF80





'

































%D







*




*







*

















              (1) Cannot be  separated  from Diphenylaraine
                                FORM  VII HCE-2
                                                                     Rev.9/88

-------
                              7HD
   HIGH CONCENTRATION EXTRACTABLE CONTINUING CALIBRATION CHECK
Lab Name:
Lab Code:
Instrument ID:
Lab File ID:
Contract :
Case No.: SAS No.:
Calibration Date:
Init. Calib. Date(s) :

SDG No . :
Time:

Min RRF80 for SPCC(#) = 0.050
Max %D for CCC(*) = 25.0%
I
| COMPOUND
i
I Chrysene
j 3 , 3 ' -Dichlorobenzidine
| bis ( 2-Ethylhexyl ) phthalate
| Di-n-octylphthalate
| Benzo (b) f luoranthene
I Benzo (k) f luoranthene
| Benz o ( a ) pyrene
| Indeno (1,2,3 -cd ) pyrene
I Dibenz (a, h) anthracene
j Benzo (g,h,i)perylene
1 Nitrobenzene-d5
j 2-Fluorobiphenyl
| Terphenyl-dl4
| Phenol -d5
I 2-Fluorophenol
1 2 , 4 , 6-Tribromophenol
i —
(LAB FILE ID:
| RRF10 -
t - - i- -
| Monochlorobiphenyl
| Dichlorobiphenyl
j Trichlorobiphenyl
| Tetrachlorobiphenyl
I Pentachlorobiphenyl
I Hexachlorobiphenyl
j Heptachlorobiphenyl
I •" ~
ILAB FILE ID:
I RRF20 =
1,
— 	
| Octachlorobiphenyl
| Nonachlorobiphenyl
| Decachlorobiphenyl
1
RRF



k


k











RRF
	 ___, 	 . 	









RRF
	




RRF80


















RRF30
^^ 	 . 	 .









RRF60





%D
















m

%D
	 	 _,









%D





                                FORM VII HCE-3
                Rev.1/89

-------
                              7HE
  HIGH CONCENTRATION CONTINUING CALIBRATION OF MULTICOMPONENT ANALYTES
Lab Name:
Lab Code :

Case No. :
Contract:
SAS No . : S
                                                        SDG No.:
Instrument ID:
                             GC Column ID:
COMPOUND
Toxaphene
Date:
Time:
Aroclor 1016
Date:
Time:
Aroclor 1248
Date:
Time:
Aroclor 1254
Date:
Time:
Aroclor 1260
Date:
Time:

AMOUNT
(ng>





PEAK
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
RT


























%D


























INITIAL
CAL. FACTOR


	























CONTINUING
CAL. FACTOR


























RPD
















=1







 * Denotes required peaks
page
of
                                FORM VII HCA
                                                            Rev. 9/88

-------
                               8 HA
    HIGH CONCENTRATION VOLATILE INTERNAL STANDARD AREA  SUMMARY
Lab Name:

Lab Code:
Case No.:
Lab File ID' (Standard) :

Instrument  ID:
Contract:

 SAS No.:
SDG No.:
                            Date Analyzed:

                            Time Analyzed:
1
1
1 _ 	 _
1 *^ —
| 12 HOUR STD
i
| UPPER LIMIT
I _________ 	
| LOWER LIMIT
1
| EPA SAMPLE
| NO.
oil
021
031
04|
05|
06|
07|
08|
09|
10|
11!
12|
131
141
151
16!
171
181
19!
201
211
22!
ISl(BCM)
AREA f





























RT

rm 	 n 	 i—-



























IS2(DFB)
AREA #





























RT

	 	

L- _i i - - -m

























IS3(CBZ)
AREA #

	



























RT





























    IS1 (BCM) = Bromochloromethane
    IS2 (DFB) = 1,4-Difluorobenzene
    IS3 (CBZ) = Chlorobenzene
                          UPPER LIMIT = + 100%
                          of internal standard area,
                          LOWER LIMIT - - 50%
                          of internal standard area.
    # Column used to flag internal standard area values with an asterisk
page 	 of
                                FORM VIII HCV
                                                Rev. 9/88

-------
                              8HB
 HIGH CONCENTRATION EXTRACTABLE INTERNAL STANDARD AREA SUMMARY
Lab Name:

Lab Code:
Case No.:
Lab File ID (Standard):

Instrument ID:-
Contract:

 SAS No.:
SDG No.:
                            Date Analyzed:

                            Time Analyzed:

12 HOUR STD

UPPER LIMIT

LOWER LIMIT

EPA SAMPLE
NO.

















T -miu-ULL.




ISl(DCB)
AREA #
"" ""




























RT





























IS2(NPT) |
AREA #





























RT
-




























IS3(ANT)
AREA %

______________________



























1
RT j
1
i
1
1

I

1
1
I
1
1
1
1
1
1
I
1
1
1
1
1
1
1
I
1
i
1
1
i
1
1
   01
   02
   03
   04
   05
   06
   07
   08
   09
   10
   11
   12
   13
   14
   15
   16
   17
   18
   19
   20
   21
   22
    IS1 (DCB) = l,4-Dichlorobenzene-d4
    IS2 (NPT) - Naphthalene-d8
    IS3 (ANT) = Acenaphthene-d8
                          UPPER LIMIT = +  100%
                          of internal standard  area.
                          LOWER LIMIT - -  50%
                          of internal standard  area.
    # Column used to flag internal standard area values  with an asterisk
page 	 of
                                 FORM VIII  HCE-1
                                                 Rev.  9/88

-------
                              8HC
 HIGH CONCENTRATION EXTRACTABLE INTERNAL STANDARD AREA  SUMMARY
Lab Name:

Lab Code:
Case No.:
Lab 'File ID  (Standard) :

Instrument ID:
Contract:

 SAS No.:
SDG No.
                            Date Analyzed:

                            Time Analyzed:

	
12 HOUR STD
— — — — —— !
	 =
UPPER LIMIT
——————————— s
LOWER LIMIT

EPA SAMPLE
NO.






•















IS4(PHN)
AREA #



SSSSSSSSSSSa _-- —























.
RT



______
























ISS(CRY)
AREA #

.

====== =====
























RT



—— ====
























IS6(PRY)
AREA $



__________
























RT



======
























   01
   02
   03
   04
   05
   06
   07
   08
   09
   10
   11
   12
   13
   14
   15
   16
   17
   18
   19
   20
   21
   22
    IS4 (PHN) - Phenanthrene-dlO
    IS5 (CRY) = Chrysene-dl2
    IS6 (PRY) = Perylene-dl2
                          UPPER LIMIT =• + 100%
                          of internal standard area.
                          LOWER LIMIT - - 50%
                          of internal standard area,
      Column used to flag internal standard area values with an asterisk
page _ of _
                                FORM VIII HCE-2
                                                Rev. 9/88

-------
                             8HD
            HIGH CONCENTRATION ANALYTICAL SEQUENCE
Lab Name:

Lab Code:
                    Case No.:
Contract:

 SAS No.:
                                                 SDG No.:
Instrument ID:
                               GC Column ID:
Date(s) Analyzed:
 THE ANALYTICAL SEQUENCE OF INITIAL CALIBRATION STANDARDS, PERFORMANCE
EVALUATION MIXES, INSTRUMENT BLANKS, METHOD BLANKS, SAMPLES, AND  CONTROL
                        MATRIX SPIKES IS GIVEN BELOW.
       01|
       02 |
       03)
       04 I
       05 |
       06 |
       07 |
       08 |
       09 |

       111
       141;
       18|
       191
       20
       211.
       22 |
       23
       24|
       25|
       26|
       27|
       28 |
       29|
       301;
       31|
       32|
       33
       34
       35|
       36|
EPA
SAMPLE NO.




































LAB
SAMPLE ID





'






























DATE
ANALYZED




































TIME
ANALYZED




































%D
TMZ |




































%D
DEC *




































page
# Column used to flag values outside of QC limits  (+ 0.5%)

   of
                                FORM VIII HCA
                                                            Rev. 9/88

-------
                              9 HA
            HIGH CONCENTRATION EXTRACTABLE GPC CALIBRATION
Lab Name:

Lab Code:
Case No.:
GPC Column ID:
Contract:.

 SAS No.:
SDG No,
                  Calibration Date:
COMPOUND
Polystyrene
Bis (2-ethylhexyl ) phthalate
Pentachlorophenol
Perylene

INITIAL]
CALIB.
RT




CALIB.
VERIF.
RT




%
D




QC
LIMITS
5%
5%
      THIS GPC CALIBRATION APPLIES TO THE  FOLLOWING  SAMPLES, AND CMS:
| EPA
| SAMPLE NO.
on
02]
03|
04|
051
06|
07|
08|
091
10|
111
121
131
14|
15|
161
171
181
191
20|
211
221
LAB
SAMPLE ID






















LAB
FILE ID






















DATE
ANALYZED











.










TIME
ANALYZED





















1
COMMENTS:
page 	 of
                                FORM IX HCE
                                                Rev.9/88

-------
                              9HB
  HIGH CONCENTRATION SINGLE COMPONENT PESTICIDE RETENTION TIMES
Lab Name:

Lab Code:
Instrument ID:
           Case No.:
Contract:

 SAS No.:
SDG No.
                             GC Column ID:
Date(s) Analyzed:
COMPOUND
alpha- BHC
beta-BHC
delta-BHC
gamma-BHC
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4,4' -DDE
Endrin
. Endosulfan II
4, 4 '-ODD
Endosulfan sulfate
4,4' -DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha-Chlordane
gamma-Chlordane

RT





















RT W]
FROM





















CNDOW
TO




















     Retention times of single component pesticides must be determined
     at least once per initial calibration on each instrument and column
     used for the analysis of Aroclors and Toxaphene.  See the Forms
     instructions for the calculation of retention time windows.  These
     data must be submitted with the data package for each group of
     samples analyzed under that initial calibration.
COMMENTS:
page
of
                                FORM IX HCA
                                                        Rev. 9/88

-------
                              10H
      HIGH CONCENTRATION AROCLOR IDENTIFICATION SUMMARY
                                         EPA SAMPLE NO.
 Lab  Name:
 Lab  Code:
Case No.:
Contract:
 SAS No.:
SDG No.:
 Instrument ID (1):
 GC  Column ID (1):
 Lab Sample ID (1):
                     Instrument ID (2):
                     GC Column ID (2):
                     Lab Sample ID (2):
COMPOUND

COLUMN 1
COLUMN 2

COLUMN 1
COLUMN 2

COLUMN 1
COLUMN 2
PEAK
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
RT










======





















RT w:
OF STJ
FROM
































ENDOW
UTDARD
TO




















"










CONCENTRATION

































MEAN
CONCENTRATION



'
'.





%D


===S=S=

* Denotes required peaks.  Report the  lower mean concentration  on  Form  I.
  If %D >25%, flag the value on  Form I according to the instructions  in Ex.  B
page 	 of 	
                                FORM X HCA
                                                Rev.  9/88

-------
                              11H
          HIGH CONCENTRATION AROCLOR DIOL CARTRIDGE CHECK
Lab Name:
Lab Code:
Case No.:
Contract:
 SAS No.:
SDG NO.:
Diol Cartridge Lot Number:
                       Date of Analysis:


COMPOUND
Aroclor 1254

SPIKE
ADDED
(ng)


SPIKE
RECOVERY
(ng)


%
REC



QC.
LIMITS

80-110

   THIS LOT OF DIOL CARTRIDGES WAS USED FOR THE FOLLOWING SAMPLES AND CMS:
COMMENTS:
| EPA
| SAMPLE NO.
on
021
031
04|
051
06 1
07 |
08 |
091
101
111
121
131
141
151
161
17|
181
191
20|
211
221
231
241
251
LAB
SAMPLE ID

























page 	 of 	
                                FORM XI HCA
                                                 Rev. 9/88

-------
                  EXHIBIT C
       TARGET COMPOUND LIST (TCL) AND
CONTRACT REQUIRED QUANTITATION LIMITS (CRQL)
                                                   Rev. 9/88

-------

-------
                        TargetCompound List  (TCP and
Contract Recruired Ouantitation Limits (CROD*
Volatiles
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
31.
32.
33.
34.
Chlorome thane
Bromome thane
Vinyl Chloride
Chloroethane
Methylene Chloride
Acetone
Carbon Disulf ide
1 , 1-Dichloroethene
1, 1-Dichloroe thane
1,2-Dichloroethene (total)
Chloroform
1 , 2-Dichloroethane
2 - Butanone
1,1, 1-Trichloroe thane
Carbon Tetrachloride
Vinyl Acetate
Bromodichl or ome thane
1,2-Dichloropropane
cis - 1 , 3 -Dichloropropene
Trichloroethene
Dibromochlorome thane
1 , 1 , 2-Trichloroethane
Benzene
trans -1,3 - Dichloropropene
Bromoform
4-Methyl-2-pentanone
2-Hexanone
Tetrachloroethene
1,1,2, 2-Tetrachloroethane
Toluene
Chlorobenzene
Ethylbenzene
Styrene
Xylene (Total)
CAS Number
74-87-3
74-83-9
75-01-4
75-00-3
75-09-2
67-64-1
75-15-0
75-35-4
75-34-3
540-59-0
67-66-3
107-06-2
78-93-3
71-55-6
56-23-5
108-05-4
75-27-4
78-87-5
10061-01-5
79-01-6
124-48-1
79-00-5
71-43-2
10061-02-6
75-25-2
108-10-1
591-78-6
127-18-4
79-34-5
108-88-3
108-90-7
100-41-4
100-42-5
1330-20-7
Quantitation Limits
(me/Re)
5.0
5.0
5.0
5.0
2.5
5.0
2.5
2.5
2.5
2.5
2.5
2.5
5.0
2.5
2.5
5.0
2.5
2.5
2.5
2.5
2.5
2.5
2.5
2.5
2.5
5.0
5.0
2.5
2.5
2.5
2.5
2.5
2.5
2.5
*Specific quantitation limits are highly matrix dependent.  The quantitation
limits listed herein are provided for guidance and may not always be
achievable.
                                   C-2
Rev. 9/88

-------
                       Target Compound List (TCLV and
Contract Reauired Ouantitation

35.
36.
37.
38.
39.
40.
41.
42.
43.

44.
45.

46.
47.
48.
49.
50.
51.
52.

53
54.
55.
56.
57.
58.

59.
60.
61.
62.
63.
64.
65.
66.
67.
68.
Extractables
Phenol
bis(2-Chloroethyl) ether
2 - Chlorophenol
1 , 3-Dichlorobenzene
1 , 4- Dichlorobenzene
Benzyl alcohol
1 , 2 -Dichlorobenzene
2-Methylphenol
bis (2-Chloroisopropyl)
ether
4 - He thy Ipheno 1
N-Nitroso-di-n-
dipropylamine
Hexachloroe thane
Nitrobenzene
Isophorone
2-Nitrophenol
2 ,4-Dimethylphenol
Benzole acid
bis(2-Chloroethoxy)
methane
2 , 4-Dichlorophenol
1,2,4- Tr ichlorobenz ene
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
4-Chloro- 3 -me thy Ipheno 1
(para-chloro-meta-cresol)
2-Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6 -Trichlorophenol
2,4, 5 -Trichlorophenol
2 - Chlor onaphthalene
2-Nitroaniline
Dimethylphthalate
Acenaphthylene
2,6-Dinitrotoluene
3-Nitroaniline
CAS Number
108-95-2
111-44-4
95-S7-8
541-73-1
106-46-7
100-51-6
95-50-1
95-48-7

108-60-1
106-44-5

621-64-7
67-72-1
98-95-3
78-59-1
88-75-5
105-67-9
' 65-85-0

111-91-1
120-83-2
120-82-1
91-20-3
106-47-8
87-68-3

59-50-7
91-57-6
77-47-4
88-06-2
95-95-4
91-58-7
88-74-4
131-11-3
208-96-8
606-20-2
99-09-2
Limits rCROU*
Quantitation Limits
(me/Kg}
20
20
20
20
20
20
20
20

20
20

20
20
20
20
20
20
100

20
20
20
20
20
20

20
20
20
20
100
20
100
20
20
20
100
(continued)
                                   C-3
Rev. 9/88

-------
      Extractables
                                                    Quantitation Limits
69 . Acenaphthene
70. 2,4-Dinitrophenol
71. 4-Nitrophenol
72. Dibenzofuran
73. 2,4-Dinitrotoluene
74 . Diethylphthalate
75. 4-Chlorophenyl-phenylether
76 . Fluorene
77. 4-Nitroaniline
78. 4,6-Dinitro-2-methylphenol
79. N-nitrosodiphenylamine (1)
80. 4-Bromophenyl-phenylether
81. alpha -BHC
82 . Hexachlorobenzene
83. beta- BHC
84 . Pentachlorophenol
85. gamma -BHC (Lindane)
86. Phenanthrene
87. Anthracene
88. delta-BHC
89 . Heptachlor
90 . Aldrin
91. Di-n-butylphthalate
92. Fluoranthene
93. Heptachlor epoxide
83-32-9
51-28-5
100-02-7
132-64-9
121-14-2
84-66-2
7005-72-3
86-73-7
100-01-6
534-52-1
86-30-6
101-55-3
319-84-6
118-74-1
319-85-7
87-86-5
58-89-9
'85-01-8
120-12-7
• 319-86-8
76-44-8
309-00-2
84-74-2
206-44-0
1024-57-3
94. Monochlorobiphenyl 27323-18-8
95 . Dichlorobiphenyl
96 . Trichlorobiphenyl
97. Tetrachlorobiphenyl
98. Pyrene
99. gamma-Chlordane
100. Endosulfan I
101. alpha -Chlordane
102. 4, 4' -DDE
103. Dieldrin
2051-60-7
2051-61-8
2051-62-9
129-00-0
5103-74-2
959-98-8
5103-71-9
72-55-9
60-57-1
104. Hexachlorobiphenyl 26601-64-9
105. Pentachlorobiphenyl 25429-29-2
20
100
100
20
20
20
20 .
20
100
100
20
20
20
20
20
100
20
20
20
20
20
20
20
20
20
100
100
100
100
20
20
20
20
20
20
100
100
(continued)
                                   C-4
                                                                   Rev. 9/88

-------
     Extractables
CAS Number
Quantisation Limits
     (me/Kg)	
106.  Endrin                       72-20-8
107.  Endosulfan II             33213-65-9
108.  4,4'-ODD         .            72-54-8
109.  Heptachlorobiphenyl       28655-71-2
110.  Butylbenzylphthalate         85-68-7

111.  Endosulfan sulfate         1031-07-8
112.  4,4'-DDT                     50-29-3
113.  Endrin ketone             53494-70-5
114.  Benzo(a)anthracene           56-55-3
115.  Methoxychlor                 72-43-5

116.  Chrysene                    218-01-9
117.  Octachlorobiphenyl        55722-26-4
118.  3,3'-Dichlorobenzidine       91-94-1
119.  bis(2-Ethylhexyl)phthalate  117-81-7
120.  Nonachlorobiphenyl        53742-07-7
121. Decachlorobiphenyl         2051-24-3
122. Di-n-octylphthalate         it?-84-0
123. Benzo(b)fluoranthene        205-99-2
124. Benzo(k)fluoranthene        207-08-9
125. Benzo(a)pyrene               50-32-8

126. Indeno(l,2,3-cd)pyrene      193-39-5
127. Dibenz(a.h)anthracene        53-70-3
128. Benzo(g,h,i)perylene        191-24-2

*Specific quantitation limits are highly matrix dependent.  The quantitation
limits listed herein are provided for guidance and may not always be
achievable.
                          20
                          20
                          20
                         100
                          20

                          20
                          20
                          20
                          20
                          20

                          20
                         200
                          40
                          20
                         200

                         200
                          20
                          20
                          20
                          20

                          20
                          20
                          20
                                    C-5
                                   Rev.  9/88

-------
                        Target Compound List (TCP and
                Contract Required Quantitation Limits  (CROP*
   Aroclor-Spec i fic/Toxaphene
     bv GC/EC Method
CAS Number
Quantitation Limits
     frog/Kg)
129.
130.
131.
132.
133.
134.
135.
136.
Toxaphene
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
8001-35-2
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
50
10
10
10
10
10
10
10
*Specific quantitation limits are highly matrix dependent.   The quantitation
limits listed herein are provided for guidance and may not always be
achievable.
                                   C-6
                                  Rev. 9/88

-------

-------
                    EXHIBIT D







            ANALYTICAL PROCEDURES FOR




HIGH CONCENTRATION VOLATILE ORGANIC WASTE SAMPLES
                                                         Rev.  9/88

-------

-------
                             TABLE OF CONTENTS



                                                                  Page No.

SECTION I   - INTRODUCTION	VOA D-l

SECTION II  - SAMPLE PREPARATION AND STORAGE  .  .  !	VOA D-4

     PART A - SAMPLE STORAGE	VOA D-5

     PART B - HIGH LEVEL METHODS FOR SCREENING
              AND ANALYSIS OF VOLATILE ORGANICS	VOA D-6

SECTION III - SCREENING OF METHANOL EXTRACTS
              FOR VOLATILES	  VOA D-12

SECTION IV  - GC/MS ANALYSIS OF, VOLATILE ORGANICS  	  VOA D-17
                                                                       Rev.  9/88

-------

-------
                                  SECTION I

                                INTRODUCTION
The samples received for high concentration analysis will consist of three
types of phases:  solids, water immiscible liquids, and water miscible
liquids.  It is also possible to receive samples that have multiple phases
such as soil, water and oil, in the same sample jar.  Because of this
possibility the samples are to be "phase separated" into their individual
phases.  The weight of each phase and the phase type are to be recorded on
appropriate data sheets and reported with the sample.

The phase separation techniques employed will vary according to the types
of sample received.  Since it is impossible to know the number and types of
phases that will be present in a sample, the choice of phase separation
techniques is left to the discretion of the analyst.  Various techniques
can be employed to separate the phases.  These include pipetting off liquid
phases (decanting should not be done),  centrifuging to remove suspended
solids, use of spatulas to remove solids (wooden tongue depressors work
well).  Whenever possible, phase separation operations should be done with
disposable glassware.  This eliminates the problem of cleaning contaminated
glassware.  The phases should be separated into glass containers with
teflon-lined screw caps.  This allows for storage and handling of the waste
in a safe manner.  Under no circumstances are samples to be homogenized to
eliminate separate phases.

Each individual phase is then taken through the procedures as a sub-sample.
The results of the analyses are to be reported for each phase of the
sample.

The analyical methods that follow are designed to analyze the organic
compounds on the Target Compound List (TCL) (See Exhibit C)..  The methods
are divided into the following sections:  sample preparation, screening and
analysis.   Sample preparation covers sample storage.  The analysis section
contains the GC/MS analytical methods for volatile organics.   The purge and
trap technique,  including related sample preparation, is included in the
analysis section because GC/MS operation and the purge and trap technique
are interrelated.
                                 VGA D-l
Rev. 9/88

-------
1.    Methodfor the Determination of Volatile (Purgeable)High
      Concentration Organic Compounds

1.1   Scope and Application

      These procedures are designed for the preparation of waste samples
      which may contain organic chemicals at a level greater than 20 mg/kg.

      The extracts and sample aliqupts prepared using these methods are
screened by gas chromatograph/flame ionization detector (GC/FID) for
volatile organic chemicals.   The results of these screens will determine
the dilution required for gas chroinatograph/mass spectrometer (GC/MS)
analysis.  The analytical scheme is summarized in flowchart form (Figure
Dl).
                                  VOA D-2                              Rev. 9/88 .

-------
                                                    Sample
               yes
          Estimate  relative volume
          of mulciple phases and
          perfora phase separation
          Procede with the analysis'
          of each phase as an
          Individual sample
                                  'Phai*"
                                 separation^
                                  required?
                             Weigh Ig sample phase
                                loco 20 ml vial
                                                1. Add 10 ml mechanol
                                                2. Add 100 pi surrogates
                                                3. Shake to disperse/
                                                  dissolve sample
                                                A. Allow to settle
                                              I Remove two 1 ml allquots 1
     Add 2 ul (10 ug) of nonane/
     dodecane to one of the allquots
                    (optional)
1)

>
p
Store one aliquot at
4*C for CC/MS analysis
Add 100 ul of oethanol
extract to 5 ml organic-
free water


Add 5 ul (250 ng)
Internal Standards
          Estimate concentration
          of major component(s)

                 I
          	3k	
          Determine aoount of
          extract  required for
          500-1000 ng of major
          components(s); (Hake
          secondary dilution
          If necessary}
                                                                           Analyze by purge
                                                                           and trap CC/MS
              Add appropriate
              amount of mechanol
              extract  to S ol of
              organic-free wa er
Add additional
met Hanoi so total
volume of mechanol
is 100 ul
5

Add 5 ul (250 ng)
Internal Standards
Compare concentra-
tion of aromatics
to that of other
major peaks
_J
                                                                          Analyze by purge
                                                                          and trap CC/MS
                                   FIGURE Dl -  FLOWCHART  FOR ANALYSIS  OF  HIGH CONCENTRATION
                                                  VOLATILE ORGANICS
                                                  VOA D-3
                                                                               Rev.  9/88

-------
          SECTION II
SAMPLE STORAGE AND PREPARATION
           VOA D-4
Rev. 9/88

-------
PART A  -  SAMPLE STORAGE

Procedures forSample Storage

     The samples must be protected from light and refrigerated at 4"C (12'C)
     from the time of receipt until analysis  or extraction.
                                 VGA D-5
Rev. 9/88

-------

PART B - HIGH LEVEL METHODS FOR SCREENING AND ANALYSIS OF VOLATILE

1.    Summary of Methods

1.1   Samples received for high concentration analysis will consist of
      three types of phases:  solids,  water immiscible liquids, and water
      miscible liquids.  It is possible to receive samples that will be
      multiple phase, such as soil, water and oil in the same sample
      container.  Multi-phase samples  are to be "phase-separated" into
      single phase units.

1.2   One gram aliquots of the samples are transferred to vials and
      extracted with methanol.  The methanol extracts are screened for
      volatile organics by GC/FID.

1.3   If organic compounds are not detected by the screen, then a 100 uL
      aliquot of the methanol extract is analyzed by purge and trap GC/MS
      for volatile organics.  If compounds are detected by the screen, the
      screening data are used to determine the amount of raethanol extract
      appropriate for GC/MS analysis.

2.    Limitations                               N

2.1   The procedure is designed to allow detection limits for screening
      purposes as low as 0.5 - 1 rag/kg for volatile organics.  If peaks are
      present based-on the GC/FID screens, a dilution of the methanol
      extract prior to GC/MS analysis  is required.  Some samples may
      contain high concentrations of chemicals that interfere with the
      analysis of other components at lower levels; the detection limits in
      those cases may be significantly higher.

2.2   These extraction and preparation procedures were developed for rapid
      and safe handling of high concentration hazardous waste samples. The
      design of the methods does not stress efficient recoveries or low
      limits of detection for all components.  Rather, the procedures were
      designed for moderate recovery and sufficient sensitivity of a broad
      spectrum of volatile organic chemicals.  The results of the analyses
      thus may reflect only a minimum of the amount of pollutants actually
      present in some samples.

3.    Interferences

3.1   Method interferences may be caused by contaminants in solvents,
      reagents, glassware, and other sample processing hardware that lead
      to discrete artifacts and/or elevated baselines in the total ion
      current profiles.  All of these materials must be routinely
      demonstrated to be free from interferences under the conditions of
      the analysis by running laboratory method blanks.  Matrix
      interferences may be caused by contaminants that are coextracted from
      the sample.  The extent of matrix interferences will vary
      considerably from source to source, depending upon the nature and
      diversity of the samples.
                                  VOA D-6                              Rev.  9/88

-------
3.2   Samples can be contaminated by diffusion of volatile organics
      (particularly fluorocarbons and methylene chloride) through the
      septum seal into the sample during storage and handling.

4.    Apparatus and Equipment

4.1   Analytical balance capable of accurately weighing 0.0001  g and a top-
      loading balance capable of weighing 0.1 g.

4.2   Glassware

      Bottles - 15 mL, screw cap, with teflon cap liner.

      Volumetric flasks - Class A with ground glass stoppers.

      Vials - 2 mL for GC autosampler.

5.    Reagent Specifications

5.1   Reagent water is defined as water in which an interferent  is not
      observed at the minimum detection limit of the parameters of
      interest.

      5.1.1   Reagent water may be generated by passing tap  water through
              a carbon filter bed containing activated carbon  (Calgon
              Corp., Filtrasorb-300,  or equivalent).

      5.1.2   A water purification system  (Millipore Super-Q or equivalent)
              may be used to generate reagent water.

      5.1.3   Reagent water may also be prepared by boiling water for 15
              minutes.  Subsequently, while maintaining the temperature at
              90°C, bubble a contaminant-free, inert- gas through the water
              for 1 hour.  While still hot, transfer the water to a narrow-
              mouth, screw-cap bottle and  seal with a teflon-lined septum
              and cap or maintain the water under a continuous purge of
              inert gas.

5.2   Methanol - Pesticide residue analysis grade,  or equivalent.   Methanol
      must be demonstrated to be free from purgeable interferences.

5.3   Preparation of Spiking Standards and Analytical Standards

      5.3.1   Stock standard solutions may be prepared from pure standard
              materials or purchased  as certified (compound purity of 96%
              or greater) solutions.  Commercially prepared stock standards
              may be used at any concentration if they are certified by the
              manufacturer or by an independent source.  Prepare stock
             . standard solutions in methanol using assayed liquids or gases
              as appropriate.

      5.3.2   Great care must be taken to maintain the integrity of all
              standard solutions.  All standard solutions should be stored
              at -10"C to -20°C in amber bottles with teflon liners in the
                                 VGA D-7
Rev. 9/88

-------
              screw caps.  For storage of calibration standards, bottle
              caps with syringe valves are recommended.

      5.3.3    Place about 9.8 mL of methanol into a 10.0 mL tared,  ground-
              glass-stoppered, volumetric flask.  Allow the flask to stand,
              unstoppered, for about 10 minutes or until all alcohol-wetted
              surfaces have dried.  Weigh the flask to the nearest
              0.0001 g.

      5.3.4    Add the assayed reference material as described below.

              5.3.4.1    Liquids - Using a 100 uL syringe,  immediately add
                         two or more drops of assayed reference material to
                         the flask, then reweigh.   The liquid must fall
                         directly into the alcohol without contacting the
                         neck of the flask.

              5.3.4.2    Gases - To prepare standards for any of the four
                         halocarbons that boil below 30°C (i.e.,
                         bromomethane, chloroethane, chloromethane,  and
                         vinyl chloride), fill a 5 mL valved, gas-tight
                         syringe with the reference standard to the 5.0 mL
                         mark.  Lower the needle to 5 mm above the methanol
                         meniscus.  Slowly introduce the reference standard
                         above the surface of the liquid.  The heavy gas
                         rapidly dissolves in the methanol.

      5.3.5    Reweigh, dilute to volume, stopper, then mix by inverting the
              flask several times.  Calculate the concentration from the
              net gain in weight.

      5.3.6    Transfer the stock standard solution into a teflon-sealed,
              screw-cap bottle.  Store with minimal headspace at -10"C to
              -20*C and protect from light.

      5.3.7    Prepare fresh stock standards every two months for the four
              gases.  All other standards must be replaced after six
              months, or sooner if comparison with check standards indicate
              a problem.

      5.3.8    Using stock standard solutions, prepare secondary dilution
              standards in methanol that contain the compounds of interest,
              either singularly or mixed together.  (See GC/MS Calibration
              in Exhibit E.)  Secondary dilution standards should be stored
              with minimal headspace and should be checked frequently for
              signs of degradation or evaporation, especially just prior to
              preparing calibration standards from them.

5.4   Surrogate Standard Spiking Solution -  Prepare stock standard
      solutions for toluene-dg,  p-bromofluorobenzene (BFB),  and 1,2-
      dichloroethane-d^ in methanol.   Prepare a surrogate standard spiking
      solution from these stock standards at a concentration of 250  ug/mL
      in methanol.
                                 VGA D-8
Rev. 9/88

-------
5.5   Purgeable Organic Control Matrix Spiking Solution - Prepare in
      methanol a spiking solution that contains the following compounds at
      a concentration of 250 ug/mL:  1,1-dichloroethene, trichloroethene,
      chlorobenzene , toluene, and benzene.

5.6   Screening Standards - Prepare standard mixture #1 containing benzene,
      toluene, ethylbenzene ,  and xylene in methanol.  Prepare standard
      mixture #2 containing n-nonane and n-dodecane in methanol.

      5.6.1    Stock  standard solutions  (1.00 mg/mL) can be prepared from
               pure standard  materials or  purchased  as certified solutions.

               5.6.1.1    Prepare stock standard solutions by accurately
                         weighing approximately 0.0100  grams of pure
                         material.  Dissolve the material in methanol;
                         dilute to volume in a 10 mL volumetric  flask.
                         Larger volumes can be used at  the convenience of
                         the analyst.  Compound purity  should be certified
                         at  96% or greater.

               5.6.1.2    Transfer the stock standard solutions  into teflon-
                         sealed, screw-cap bottles.  Store with minimal
                         headspace at -10'C to -20'C and protect from
                         light.  Stock standard solutions should be checked
                         frequently for signs of degradation or
                         evaporation., especially just prior to preparing
                        . calibration standards from them.
      5.6.2    Prepare  working  standards of mixtures #1 and #2 at 10 ng/uL
               of each  compound in methanol.  Prepare spiking mixture of #2
               at 5000  Mg/mL in methanol.  The spiking mixture is added to
               each  screening extract.

5.7   Internal Standard Spiking Solution - Prepare stock standard solutions
      for bromochlorome thane, 1,4-difluorobenzene, and chlorobenzene -dj in
      methanol.  Prepare an internal standard spiking solution from these
      stock standards at a concentration of 50  ug/mL in methanol.   This 50
      ug/mL solution is the working standard solution.

5.8   p-bromofluorobenzene (BFB) Standard -  Prepare a 25 ug/mL solution of
      BFB in methanol.

5.9   Calibration Standards - Prepare a stock standard solution(s)  for the
      34 volatile target compounds  listed in Exhibit C.   Prepare working
      standard solutions from these stock standards at a concentration of
      50 ug/mL in methanol.  The working standard(s)  will be  used to
      prepare calibration standards at five  specified concentrations as
      described in Section IV,  6.3.1.  Prepare  fresh working  standards
      weekly for the four gases. All other  working standards must  be
      replaced after one month.
                                 VOA D-9
Rev. 9/88

-------
6.2
6.3
6.4
7.
 Phase Separation

6.1    The samples received for high concentration analysis will
       consist of three phase types:

       Water miscible liquid
       Water immiscible liquid
       Solid

 It is possible to receive samples that contain multiple phases such
 as  water, oil, and soil in the same sample jar.  Because of this
 possibility, the samples are to be "phase separated" into individual
 phases.

 Each individual phase is taken through the procedures as a subsample.
 Report analytical results for each sample phase.

 Do not analyze any phase that represents less than 10% of the total
 sample volume.

 In the following procedures, where applicable, references to
 "samples" explicitly mean "single phase units".

 Sample Preparation for Volatile Oreanics
      The high concentration method is based on extracting the sample with
      methanol.  A portion of the extract is used for a screen on the
      GC/FID.  From the results of the screen, an aliquot of the methanol
      extract is added to reagent water containing the internal standards.
      This is purged at ambient temperature,.

7.1   Weigh 1 g of each sample phase into separate tared 15 mL vials using
      a top loading balance.  Record the actual weight to the nearest
      0-1 g.

7.2   Weigh 1.0 g of corn oil into a tared 15 mL vial for use as the
      control matrix spike.

7.3   Quickly add 10 mL of methanol to all vials, then add 100 uL of
      surrogate spiking solution (5.4).  For the control matrix spike, add
      10 mL of methanol, 100 uL of surrogate spiking solution, and 100 uL
      of matrix spiking solution (5.5).

7.4   Prepare a method blank by adding 100 uL of surrogate spiking
      solution to 10 mL of methanol in a 15 mL vial.

7.5   Cap all vials and shake for 2 minutes.  (Note:   Steps 7.1 through 7.4
      must be performed rapidly to avoid loss of volatile organics.   These
      steps must be performed in a laboratory free of solvent fumes.)

7.6   In order to enhance sample extraction with methanol,  the sample may
      be agitated in cold water in an ultrasonic bath for 2-3 minutes.
                                 VOA D-10
                                                                  Rev.  9/88

-------
7.7   Transfer for storage 1 mL of extract to a GC vial using a disposable
      pipet.  Transfer an additional 1 mL of extract to a GC vial for use
      as a screening extract.  These extracts may be stored in the dark at
      4°C prior to analysis.  Mark the level of liquid in the vial to check
      for evaporation of the extract.   Add 2 uL of 5000 ug/mL #2 spiking
      mixture (n-nonane and n-dodecane) (5.6.2) to each screening extract.
      The sample is now ready for screening.
                                 VOA D-ll
Rev. 9/88

-------
         SECTION III
SCREENING OF METHANOL EXTRACTS
           FOR VOLATILES
          VOA D-12
Rev. 9/88

-------
1.    Summary of Method

      The methanol extracts of the samples are screened on a gas
      chromatograph/flame ionization detector (GC/FID).   The results of the
      screen will determine at what dilution level the volatile organics
      extracts are to be analyzed by GC/MS.

2.    Apparatus and Materials

2.1   Gas Chromatograph - A gas chromatograph suitable for on-column
      injection, and all required accessories including syringes,
      analytical columns, gases,  flame ionization detector,  and strip-chart
      recorder.  A data system is recommended for measuring peak areas.

2.2   GC Column - 3 m x 2 mm ID glass column packed with 10% SP-2100 on
      100-120 mesh Supelcoport (or equivalent).   The column temperature
      should be programmed from 55"C to 280'C at 16°C/minute and held at
      280°C for 10 minutes.  (Other chromatographic conditions may also
      provide acceptable results.)

3.    Reagents

3.1   Methanol - Pesticide residue analysis  grade,  or equivalent.   Methanol
      must be demonstrated to be  free from purgeable interferences.

3.-2   Screening Standards - Prepare standard mixture #1 containing benzene:
      toluene, ethylbenzene,  and  xylene in methanol.   Prepare standard
      mixture #2 containing n-nonane and n-dodecane in methanol.

      3.2.1    Stock standard solutions  (1.00 mg/mL) can be prepared-from
               pure standard materials or purchased as certified solutions.

               3.2.1.1    Prepare stock standard solutions by accurately
                         weighing approximately 0.0100 grams of pure
                         material.  Dissolve the material in methanol;
                         dilute to volume in a 10 mL volumetric flask.
                         Larger volumes can be used at the convenience of
                         the analyst.  Compound purity should be certified
                         at 96% or greater.

               3.2.1.2    Transfer the stock standard solutions into Teflon-
                         sealed,  screw-cap bottles.  Store with minimal
                         headspace at -10°C  to -20°C and protect from
                         light.  Stock standard solutions should be checked
                         frequently for signs of degradation or
                         evaporation, especially just prior to preparing
                         calibration standards from them.
                                VOA D-13
Rev. 9/88

-------
      3.2.2   Prepare working standards of mixtures #1 and #2 at 10 ng/uL
              of each compound in methanol.
              Screening Standard #1
              Screening Standard #2
              Benzene                              n-nonane
              Toluene                              n-dodecane
              Ethylbenzene
              Xylene

4.    Limitations

      The flame ionization detector varies considerably in sensitivity when
      comparing aromatics and halogenated methanes and ethanes.
      Halomethanes are approximately 20 times less sensitive than aromatics
      and haloethanes are approximately 10 times less sensitive.

5.    Extract Screening

5.1   External Standard Calibration - Standardize the GC/FID each 12 hour
      shift for half scale response when injecting 1-5 uL of the screening
      standards, mix #1 (aromatics) and the mix #2 (n-nonane, n-dodecane),
      at the 10ng/pL concentration.

5.2   Inject the same volume of methanol sample extract as the standard
      mixture.

6.    Interpretation of Screening Results

6.1   Compare the methanol sample extract chromatogram with the method
      blank and standard chromatograms.

      6.1.1   If no peaks are noted, other than those also in  the method
              blank, analyze 100 uL of  the sample by GC/MS.  (See Table 1)

      6.1.2   If peaks are present prior to  the n-dodecane,  and the
              aromatics are distinguishable,  follow Option A below to
              determine the dilution needed  to analyze by GC/MS.

      6.1.3   If peaks are present prior to  the n-dodecane,  but the
              aromatics are absent or indistinguishable, use Options B or C
              below.  Calculate a factor using Equation Dl:
               Equation Dl

                        X Factor
peak area of sample major peak
  peak area of n-nonane
               See Table  1 to determine  the dilution needed to  analyze by
               GC/MS.
                                 VOA D-14
                                  Rev.  9/88

-------
6.2   Following are three options for interpreting the GC/FID chromatogram.

      6,2.1   Option A  - Use standard mixture #1 as an external standard to
              calculate an approximate concentration of the aromatics in
              the  sample.  If aromatics appear to be the most concentrated
              materials in the sample, use the screening information to
              determine the proper dilution for purge and trap.  Use a
              volume of methanol to give 500 - 1000 ng of the most
              concentrated aromatics.  This should be the best approach;
              however,  the aromatics may be absent or obscured by higher
              concentrations of other purgeables.  If this is the case,
              Options B or C may be more suitable.

      6.2.2   Option B  -  Use standard mixture #2 as an internal standard
              to calculate the factor in Equation Dl.  Use this factor and
              Table 1 to determine a dilution for purge and trap.  All
              purgeables of interest have retention times less than n-
              dodecane.

      6.2.3   Option C  - Use standard mixture #2 as an internal standard to
              estimate  the concentration of the major peaks.  Calculate a
              volume of methanol that gives 500 - 1000 ng of the most
              concentrated constituents. Use that volume of methanol for
              purge and trap analysis.

      6.2.4   If the screening results indicate that less than 5 uL of
              extract should be used for GC/MS analysis, prepare a
              secondary dilution in methanol and use the secondary dilution
              for  GC/MS analysis.

      6.2.5   NOTE:  The screening procedure can also provide information
              regarding certain late-eluting compounds (e.g., naphthalene)
              which may complicate subsequent GC/MS analyses.  Calculate
              the  appropriate dilution factor for the concentrations
              exceeding the table.
            TABLE 1 - DETERMINATION OF GC/MS PURGE AND TRAP DILUTION LEVEL
                X Factor
  Estimated
Concentration
Range (mg/kg>*
  Volume of
  Methanol
Extract (uL)**
0
0.5 -
2.5 -
12.5 -


5
10
50
250


0
1
5
25


- 10
- 20
- 100
- 500


100
50
10
100
of 1/50
dilution***
                                 VOA D-15
                                   Rev.  9/88

-------
Actual concentration ranges could be 10 to 20 times higher
than this if the compounds are halogenated and the estimates
are from a GC/FID.

The volume of methanol extract added to the 5 mL of water
being purged.  Add additional PURE methanol to bring the
total volume of methanol to 100 uL.

Concentrations at this and higher levels require an initial
dilution of the extract in methanol before removing an
aliquot for analysis by GC/MS.
                   VGA D-16
Rev. 9/88

-------
             SECTION  IV
GC/MS ANALYSIS OF VOLATILE ORGANICS
            VOA D-17
Rev. 9/88

-------
1.     Summary: ofLMethod

      A portion of the methanol extract,  prepared as described in the
      preceeding sections,  is diluted to  5 mL with reagent water.  An inert
      gas is bubbled through this solution in a specifically designed
      purging chamber at ambient temperature.   The purgeables are
      effectively transferred from the aqueous phase to the vapor phase.
      The vapor is swept through a sorbent column where the purgeables are
      trapped.  After purging is completed, the sorbent column is heated
      and backflushed with the inert gas  to desorb the purgeables onto a
      gas chromatographic column.  The gas chromatograph is temperature
      programmed to separate the purgeables which are then detected with a
      mass spectrometer.

2.     Interferences

2.1   Impurities in the purge gas, organic compounds out-gassing from the
      plumbing ahead of the trap, and solvent vapors in the laboratory
      account for the majority of contamination problems.  The analytical
      system must be demonstrated to be free from contamination under the
      conditions of the analysis by running laboratory method blanks.
      Teflon tubing, thread sealants, and flow controllers should be used
      in the purging device.

2-.2   Samples can be contaminated by diffusion of volatile organics
      (particularly fluorocarbons, acetone,  and methylene chloride)
      through the septum seal into the sample during storage and handling.
      Blanks prepared from methanol and carried through the holding period
      and the analysis protocol may serve as a check on such contamination.

2.3   Contamination by carry-over can occur whenever highly concentrated
      and minimally concentrated samples  are sequentially analyzed.  To
      reduce carry-over, the purging device and sampling syringe must be
      rinsed with reagent water between sample analyses.  Whenever a highly
      concentrated sample is encountered, it should be followed by an
      analysis of reagent water to check for cross contamination.  For
      samples containing large amounts of water-soluble materials,
      suspended solids, high boiling compounds, or high levels of
      purgeables  , it may be necessary to wash out the purging device with
      a detergent solution, rinse it with distilled water, and then dry it
      in a 105°C oven between analyses.  The trap and other parts of the
      system are also subject to contamination; therefore, frequent bakeout
      and purging of the entire system may be required.
          •
2.4   Contamination can also occur when samples contain certain late-
      eluting compounds (e.g., naphthalene).  The screening results should
      provide an indication of the presence of such contamination.  When
      contamination due to late-eluting compounds is suspected, a prolonged
      bakeout of the GC column and the trap is recommended.

3.     Apparatus. Materials. and Equipment

3.1   Micro syringes -  10 uL and larger,  0.006 inch ID needle.
                                 VGA D-18
Rev. 9/88

-------
3.2   Syringe valve  - two-way, with Luer ends (three each), if applicable
      to the purging device.

3.3   Syringe - 5 mL, gas tight, with shut-off valve.

3.4   Analytical balance, capable of accurately weighing 0.0001 g. and a
      top-loading balance capable of weighing 0.1 g.

3.5   Glassware

      3.5.1   Bottles  -  15  mL,  screw cap,  with Teflon  cap  liner.

      3.5.2   Volumetric flasks  -  class A  with ground-glass stoppers.

      3.5.3   Vials  -  2  mL  for  GC  autosampler.

3.6   Purge and Trap Device.  The purge and trap device consists of three
      separate pieces of equipment; the sample purger,  the trap, and the
      trap heater.  Several complete devices are now commercially
      available.

      3.6.1   The  sample purger  must be designed  to  accept 5 mL samples
              with a water  column  at least 3 cm deep.  The gaseous head
              space  between the  water column and  the trap  must have a  total
              volume of  less  than  15 mL.   The purge  gas must pass through
              the  water  column.as  finely divided  bubbles with a diameter of
              ' less than  3 mm  at  the origin.  The  purge gas must be
              introduced no more than 5 mm from the base of the water
              column.  The  sample  purger,  illustrated  in Figure D2, meets
              these  design  criteria.  Alternate sample purge devices may be
              utilized provided  equivalent performance is  demonstrated.

      3.6.2   The  trap must be  at  least 25 cm long and have an inside
              diameter of at  least 0.105 inch.  The  trap must be packed to
              contain the following minimum lengths of absorbents:  15 cm
              of 2,6-diphenylene oxide polymer (Tenax-GC 60/80 mesh) and 8
              cm of  silica  gel  (Davison Chemical, 35/60 mesh, grade 15, or
              equivalent).  The  minimum specifications for the trap are
              illustrated in  Figure D3.

      3.6.3   The  trap heater should be capable of rapidly heating the trap
              to 180°C.  The  polymer section of the  trap should not be
              heated higher than 180"C and the remaining sections should
              not exceed 220"C.  The heater design,  illustrated in Figure
              D3, meets  these criteria.

      3.6.4   The purge  and trap device may be assembled as a separate unit
              or be  coupled to a gas chromatograph as  illustrated in
              Figures D4 and  D5.

3.7   GC/MS System

      3.7.1   Gas Chromatograph  - A temperature programmable gas
              chromatograph suitable for on-column injection, and all

                                 VOA D-19                              Rev. 9/88

-------

        required accessories including syringes, analytical columns,
        and gases.

3.7.2   Column  - 6 ft long x 0.1 in ID glass, packed with 1% SP-1000
        on Carbopack B  (60/80 mesh) or equivalent.  NOTE:  Capillary
        columns may be  used for analysis of volatiles, as long as  the
        Contractor uses the instrumental parameters in EPA Method
        524.2 as guidelines, uses the internal standards and
        surrogates specified in this contract, and demonstrates  that
        the analysis meets all of the performance and QA/QC criteria
        contained in this contract.

3.7.3   Mass Spectrometer - Capable of scanning from 35 to 260 amu
        every seven seconds or less, utilizing 70 volts (nominal)
        electron energy in the electron impact ionization mode and
        producing a mass spectrum which meets all the criteria in
        Table 2 when 50 ng of 4-bromofluorobenzene (BFB) is injected
        through the gas chromatograph inlet.

3.7.4   GC/MS Interface - Any gas chromatograph to mass spectrometer
        interface that  gives acceptable calibration points at 50 ng
        or less per injection for each of the parameters of interest
        and achieves all acceptable performance criteria (Exhibit  E)
        may be used.  Gas chromatograph to mass spectrometer
        interfaces constructed of all-glass or glass-lined materials
        are recommended.  Glass can be deactivated by silanizing with
        dichlorodimethylsilane.

3.7.5   Data System - A computer system must be interfaced to the
        mass spectrometer that allows the continuous acquisition and
        storage on machine-readable media of all mass spectra
        obtained throughout the duration of the chromatographic
        program.  The computer must have software that allows
        searching any GC/MS data file for ions of a specified mass
        and plotting such ion abundances versus time or scan number.
        This type of plot is defined as an Extracted Ion Current
        Profile (EICP).  Software must also be available that allows
        integrating the abundance in any EICP between specified  time
        or scan number  limits.
                           VOA D-20
Rev. 9/88

-------
           Option*/
           foem
           Trip
                           14mm
                              O.D.
                         Inlet'/. in.
                              00
                         ^ Simple Inlet
                         I— 2-** ft Syringe
                         _— f 7cm 20 gmuge tyring* needle

                            6mm O.D. Rubber Septum
         10mm gtett frit
         medium porosity
                               _ 10mm O.D.   '/
                                              tt
                                                   O.D.
fa— Inlet
'/tin OO
5
\
^
4
/Steinlett Steel
3 13X molecuter
% sieve purge
yj get drier
^»_ Purge get
k //our control
          FIGURE D2    -  PURGING DEVICE
       feeling procedure
                                 Cerurtructton
    wool Smm
Crede 75
           (f
 Tone* 15cm
% OV-f
Gtett  Smen
wool
           •, f
                 inlet


           FIGURE D3
                                            Comprettion Imine,
                                            •nul end ferrule*

                                             14ft 7^foot retittenee
                                             wire wrepped tofid
                                             Thermocouple/controller
                                             tensor
                                              Tubing 25 em.
                                              O.JOS in. I.D.
                                              0 125 in. O.O
                                              Steinlett tteel
                            TRAP PACKINGS AND CONSTRUCTION TO  INCLUDE
                             DESORB CAPABILITY
                             VOA D-21
                                                                           Rev.  9/88

-------
     Corrttr got fhw contra/
Prtuurm rmgulotor
   Purgm
   How control
                              Liquid infection pong
                                              -— Confirmatory
                                                T o dittaor
                                                  Hots:
                                                    All Unit
                                                    trip ind GC
                                                    thoulti o»
                                                    to 80*C
      FIGURE DA   -  SCHEMATIC OF  PURGE AND  TRAP DEVICE  - PURGE MODE
          ' g** flow

Prmtsurt rtgutttor
   Purg* gts    	
   flow control j.
 /3X mol+culir
 
-------
4.    Reagents

4.1   Reagent water is defined as water in which an interferent is not
      observed at the minimum detection limit of the parameters of
      interest.

      4.1.1    Reagent water may be generated by passing  tap water  through a
               carbon filter bed containing activated carbon (Calgon Corp.,
               Filtrasorb-300, or equivalent).

      4.1.2    A water purification system {Millipore Super-Q or equivalent)
               may be used  to generate reagent water.

      4.1.3    Reagent water may also be prepared by boiling water  for  15
               minutes.  .Subsequently, while maintaining  the temperature at
               90*C, bubble a contaminant-free inert gas  through the water
               for 1 hour.  While still hot, transfer the water to  a narrow-
               mouth, screw-cap bottle and seal with a Teflon-lined septum
               and cap or maintain the water under a continuous purge of
               inert gas.

4.2   Methanol -  Pesticide residue analysis  grade,  or equivalent.   Methanol
      must be demonstrated to be free from purgeable interferences.

4,3   Preparation of Spiking Standards and Analytical Standards.

      4.3.1    Stock standard solutions may be prepared from pure standard
               materials or purchased as certified (compound purity of  96%
               or greater)  solutions.  Prepare stock standard solutions in
               methanol using assayed liquids or gases as appropriate.
               Commercially prepared stock standards may be used at any
               concentration if they are certified by the manufacturer or by
               an independent source.

      4.3.2    Great care must be taken to maintain the integrity of all
               standard solutions.  All standard solutions should be stored
               at -10SC to  -20°C in amber bottles with teflon liners in the
               screw caps.  For storage of calibration standards, bottle
               caps with syringe valves are recommended.

      4.3.3    Place about  9.8 mL of methanol into a 10.0 mL tared, ground-
               glass-stoppered, volumetric flask.  Allow  the flask  to stand,
               unstoppered, for about 10 minutes or until all.alcohol-wetted
               surfaces have dried.  Weigh the flask to the nearest
               0.0001 g.

      4.3.4    Add the assayed reference material as described below.

               4.3.4.1    Liquids - Using a 100 uL syringe,  immediately add
                         two or more drops of assayed reference material to
                         the flask,  then reweigh.   The liquid must fall
                         directly into the alcohol without contacting the
                         neck of the flask.
                                 VOA D-23             N                Rev.  9/88

-------
              4.3.4.2    Gases - To prepare standards for any of the four
                         halocarbons that boil below 30°C (i.e.,
                         bromomethane, chloroethane, chloromethane, and
                         vinyl chloride), fill a 5 mL valved, gas-tight
                         syringe with the reference standard to the 5.0 mL
                         mark.  Lower the needle to 5 nun above the methanol
                         meniscus.  Slowly introduce the reference standard
                         above the surface of the liquid.  The heavy gas
                         rapidly dissolves in the methanol.

      4.3.5   Reweigh, dilute to volume, stopper, then mix by inverting the
              flask several times.  Calculate the concentration from the
              net gain in weight.

      4.3.6   Transfer the stock standard solution  into a Teflon-sealed,
              screw-cap bottle.  Store with minimal headspace at -10°C to -
              20"C and protect from light.

      4.3.7   Prepare fresh stock standards every two months for the four
              gases.  All other standards must be replaced after six
              months, or sooner if comparison with  check standards indicate
              a problem.

      4.3.8   Using stock standard solutions, prepare secondary dilution
              standards in methanol that contain the compounds of -interest,
              either singularly or mixed together.  (See GC/MS Calibration
              in Exhibit E.)  Secondary dilution standards should be stored
              with minimal headspace and should be  checked frequently for
              signs of degradation or evaporation,  especially just prior to
              preparing calibration standards from  them.

4.4   Surrogate Standard Spiking Solution -  Prepare stock standard
      solutions for toluene-dg,  p-bromofluorobenzene (BFB),  and 1,2-
      dichloroethane-d^ in methanol.   Prepare a surrogate standard spiking
      solution from these stock standards at a concentration of 250 ug/mL
      in methanol.

4.5   Purgeable Organic Control Matrix Spiking Solution - Prepare in
      methanol a spiking solution that contains the following compounds  at
      a concentration of 250 ug/mL:   1,1-dichloroethene,  trichloroethene,
      chlorobenzene, toluene,  and benzene.

4.6   Internal Standard Spiking Solution -  Prepare stock standard solutions
      for bromochloromethane,  1,4-difluorobenzene,  and chlorobenzene-de  in*
      methanol.   Prepare an internal standard spiking solution from these
      stock standards at a concentration of 50 ug/mL in methanol.   This  50
      ug/mL solution is the working standard solution.

4.7   P-bromofluorobenzene (BFB) Standard -  Prepare a 25 ug/mL solution  of
      BFB in methanol.

4.8   Calibration Standards -  Prepare a stock standard solution(s)  for the
      34 target compounds listed in Exhibit C.  Prepare working standard
      solutions from these stock standards at a concentration of 50 ug/mL

                                 VOA D-24                             Rev.  9/88

-------
5.

5.1
5.3
6.

6.1
6.2
 in methanol.  The working standard(s) will be used to prepare
 calibration standards at five specified concentrations as described
 in 6.3.1.  Prepare fresh working standards weekly for the four gases.
 All other working standards must be replaced after one month.

 Instrument Operating Conditions

 The following are recommended purge and trap conditions:
            Purge Flow:

            Desorb Flow:

            Purge:


            Desorb:

            Bake:
                        25-40 mL per minute.

                        20-60 mL per minute.

                        11 minutes with  trap temperature of
                        26°C or less.

                        4 minutes at 180°C.

                        12 - 15 minutes  at 180°C.
5.2   The following are recommended operating conditions for the gas
      chromatograph:

            Column:
            Carrier:

            Temperature Program:
                              Carbopack B  (60/80 mesh) with 1%
                              SP-1000  packed in a 6 foot by
                              2 mm ID glass column.

                              Helium - 30 mL per minute.

                              45"C for 3 minutes,
                              program at 8°C per minute to
                              220eC and hold for 15 minutes.
Establish the following operating conditions for the mass
spectrometer:
            Electron Energy:

            Mass Range:

            Scan Time:
                              70 Volts (nominal)

                              35 - 260

                              To give at least 5 scans per peak,
                              but not. to exceed 3 seconds per
                              scan.
Calibration
Assemble a purge and trap device that meets the specification in
paragraph 3.6.  Condition the trap overnight at 180°C in the purge
mode with an inert gas flow of at least 20 mL/min.  Prior to use,
daily condition traps 10 minutes while backflushing at 180°C with the
column at 220°C.

Connect the purge and trap device to a gas chromatograph.   The gas
chromatograph must be operated using temperature and flow rate
                                 VOA D-25
                                                                 Rev.  9/88

-------
r
                     parameters equivalent to those In paragraph 5.2. Calibrate the purge
                     and trap-GC/MS system using the internal standard technique described
                     below.

               6.3   Internal Standard Calibration Procedure.  The three internal
                     standards are:  bromochloromethane, 1,4-difluorobenzene, and
                     chlorobenzene-dj.

                     6.3.1    From working standards  in  methanol, prepare calibration
                              standards at the following five concentration  levels  for  each
                              target parameter:  20,  50,  100, 150, and 200 ug/L  in  reagent
                              water (corresponding to 10, 25, 50, 75, and 100 rag/kg in  a  1
                              g sample).   Add additional methanol so  that the total amount
                              of methanol  is 100 uL per  5 mL of aqueous standards.   Aqueous
                              standards may be stored up to 24 hours, if held in sealed
                              vials with zero headspace.  If not so stored,  they must be  *
                              discarded after an hour unless they are set up to  be  purged
                              by an autosampler.  When using an autosampler,  the standards
                              may  be kept  up to 12 hours in purge tubes connected via the
                              autosampler  to the purge and trap device.

                     6.3.2    Add  5 uL of  the 50 ug/mL internal standard spiking solution
                              (4.6) to each 5 mL standard, blank, and sample.

                     6.3.3  •  After the ion abundance criteria for BFB (Table 2) are met,
                              analyze each calibration standard (6.1.3).  Tabulate  the  area
                              response of  the characteristic  ions against the concentration
                              for  each compound and internal standard.  Calculate relative
                              response factors (.RRF)  for each compound using Equation D2.
                              The  characteristic ions and internal standards for the
                              volatile compounds are  shown in Table 3.
                           TABLE 2 - BFB KEY IONS AND ABUNDANCE CRITERIA
                           Mass
Ion Abundance Criteria
                           50
                           75

                           95
                           96

                           173

                           174

                           175

                           176


                           177
15.0 - 40.0 percent of mass 95

30.0 - 60.0 percent of mass 95

Base peak, 100 percent relative abundance
5.0 - 9.0 percent of mass 95
Less than 2.00 percent of mass 95

Greater than 50.0 percent of mass 95

5.0 - 9.0 percent of mass 174

Greater than 95.0 percent but less than 101.0
percent, of mass 174

5.0 - 9.0 percent of mass 176
                                                VOA D-26 .
                                         Rev.  9/88

-------
NOTE 1:  BFB  criteria MUST  be  met before any  samples, blanks,  or
         standards  are  analyzed.  Any  samples analyzed when  tuning
         criteria have  not  been met will require  reanalysis  at no
         additional cost  to the Agency.

NOTE 2:  Whenever the laboratory takes corrective action which may
         change  or  affect the  tuning criteria for BFB (e.g.,  ion
         source  cleaning  or repair, etc.),  the tune must be  verified
         irrespective of  the 12-hour tuning requirements.

NOTE 3:  The  twelve (12)  hour  time period  for GC/MS system tuning  and
         standards  calibration (initial or  continuing calibration
         criteria)  begins at the moment of  injection of the  BFB
         analysis that  the  laboratory  submits as  documentation of
         compliant  tune.  The  time period ends after twelve  (12) hours
         has  elapsed according to the  system  clock.
Equation D2  RRF  -
                                         X  Cis
                                               x
        Where:
                   Area of the characteristic ion for
                   the compound to be measured.

                   Area of the characteristic ion for
                   the specific internal standard from
                   Table 3.
        Cis -
    Concentration of the internal
    standard.
        GX  -      Concentration of the compound to be
                   measured.
                           VGA D-27
                                                 Rev. 9/88

-------
TABLE 3 - CHARACTERISTIC IONS FOR VOLATILE COMPOUNDS
Compound
SURROGATE STANDARDS
4- Bromof luorobenzene
1 , 2-Dichloroethane-d^
Toluene-dg
INTERNAL STANDARDS
Bromochlorome thane (1)
1,4-Dif luorobenzene (2)
Chlorobenzene-dj (3)
TARGET COMPOUNDS
Chlorome thane
Bromorae thane
Vinyl Chloride
Chloroe thane
Methylene Chloride
Acetone
Carbon Disulfide
1,1-Dichloroe thene
1,1-Dichloroe thane
1,2- Dichloroethene
Chloroform
1 , 2 -Dichloroe thane
2-Butanone
1,1,1 -Trichloroe thane
Carbon Tetrachloride
Vinyl Acetate
Bromodichlorome thane
1 , 2-Dichloropropane
trans - 1 , 3-Dichloropropene
Trichloroe thene
Dibromochlorome thane
1,1,2 -Trichloroe thane
Benzene
c is - 1 , 3-Dichloropropene
Bromoform
2-Hexanone
4 -Methyl - 2 - Pentanone
Tetrachloroe thene
1,1,2, 2-Tetrachloroethane
Toluene
Chlorobenzene
Ethylbenzene
Styrene
Total Xylenes
Primary
Ion

95
65
98

128
114
117

50
94
62
64
84
43
76
96
63
96
83
- 62
72
97-
117
43
83
63
75
130
129
97
78
75
173
43
43
164
83
92
112
106
104
106
Secondary Internal
Ion(s) Standard

174, 176
102
70, 100

49, 130, 51
63, 88
82, 119

52
96
64
66
49, 51, 86
58
78
61, 98
65, 83, 85, 98,. 100
61, 98
85
64, 100, 98
57
99, 117, 119
119, 121
86
85, 129
65, 114
77
95, 97, 132
208, 206
83, 85, 99, 132, 134
	
77
171, 175, 250, 252, 254, 256
58, 57, 100
58, 100
129, 131, 166
85, 131, 133, 166
91
114
91
78, 103
91

3
1
3

1
2
3

1
1
1
1
1
1
1
. 1
1
1.
1
1
1
2
2
2
2
2
2
2
2
2
2
2
" 2
3
3
3
3
3
3
3
3
3
                     VOA D-28
Rev. 9/88

-------
      6.3.4    The average relative response factor (RRF)  must be calculated
               for all  compounds.   A system performance  check must be made
               before this calibration curve is  used.  The following five
               System Performance  Check Compounds  (SPCC) are checked for a
               minimum  average  response factor:  chlororaethane,  1,1-
               dichloroethane,  bromoform,  1,1,2,2 -tetrachloroethane, and
               chlorobenzene.   The minimum average  response factor for
               bromoform is 0.250;  the minimum average response factor for
               all other SPCCs  is  0.300.

               The following  five  Calibration Check Compounds (CCC) are used
               to  evaluate the  curve:   vinyl chloride, 1,1-dichloroethene,
               chloroform,  1,2-dichloropropane,  toluene, and ethylbenzene.
               Calculate the  Percent Relative Standard Deviation <%RSD)
               (Exhibit E,  Section III,  Equation 2.2) of RRF values over the
               working  range  of the curve.   A maximum %RSD of 30.0% for each
               CCC must be met  before the  curve  is  valid.

      6.3.5    Once each 12 hours  (beginning with  the injection of the
               tuning compound), after demonstrating the key ion abundance
               criteria for BFB (Table 2),  a continuing  calibration check
               must be  performed.   Analyze  the 50 ug/L standard,  using the
               same conditions  as  those used for the initial calibration.

               6.3.5.1     Calculate  the  response factors as  in 6.3.3.   Check
                          the minimum  response factors for  the system
                         performance  check compounds (SPCC). . The  minimum
                          response  factor for bromoform is 0.250;  the
                         minimum response factor for all other  SPCCs is
                         0.300.

               6.3.5.2    Calculate  the response factors for  the  calibration
                         check compounds (CCC)  and  compare  against those
                         determined in  the  initial  calibration.   Calculate
                         the percent  difference (%D) (Exhibit E,  Section
                         III,  Equation 2.3).  The maximum  %D for CCCs  is
                         25.0%.

               6.3.5.3    After verifying that the continuing calibration
                         meets the  criteria  for SPCCs and CCCs,  calculate
                          the relative  response  factors for  the  remaining
                         compounds.

7.    Sample Analysis

      An aliquot of the methanol extract is added to reagent water
      containing the internal  standards.  This is  purged at  ambient
      temperature and analyzed by GC/MS.

7.1   All samples and standard solutions must be allowed to warm to  ambient
      temperature before analysis.

7,2   Adjust the purge gas (helium) flow rate to 25-40 (±3) mL per minute.
      Variations from this flow rate may be necessary to achieve better

                                 VGA D-29                              Rev.  9/88

-------
      purging and collection efficiencies for some compounds, particularly
      chloromethane and broraoform.

7.3   After achieving the key ion abundance criteria for BFB (Table 2),
      perform the daily system calibration, as described in 6.3.5.

7.4   The purgeable organics screening procedure will show the approximate
      concentrations of major sample components.  If a dilution of the
      sample is indicated, this dilution shall be made just prior to GC/MS
      analysis of the sample.

7.5   Remove the plunger from a 5 mL "Luerlock" type syringe equipped with
      a syringe valve and fill until overflowing with reagent water.
      Replace the plunger and compress the water to vent trapped air.
      Adjust the volume to 4.9 mL.   Full the plunger back to 5 mL to allow
      volume for the addition of sample and standards.   Add 5 uL of the
      internal standard solution.  Add the volume of methanol extract
      determined from the screening procedure and a volume of pure methanol
      to total 100 uL (excluding methanol in the standards).

7.6   Attach the syringe/syringe valve assembly to the syringe valve on the
      purging device.  Open the syringe valve and inj ect the water/
      methanol sample into the purging chamber.

7.7   Proceed with the analysis as outlined.  Analyze all method blanks on
      the same instrument as the samples.

7.8   Close both valves and purge the sample for 11.0 (±0.1) .minutes at
      ambient temperature.      •  „

7.9   At the conclusion of the purge time, attach the trap to the
      chromatograph, adjust the device to the desorb mode, and begin the
      gas chromatographic temperature program.  Concurrently, introduce the
      trapped materials to the gas chromatographic column by rapidly
      heating the trap to 180°C while backflushing the trap with an inert
      gas between 20 and 60 mL/min for 4 minutes. If this rapid heating
      requirement cannot be met, the gas chromatographic column must be
      used as a secondary trap by cooling it to 30°C (or subambient, if
      problems persist) instead of the recommended initial temperature of
      45'C.

7.10  While the trap is being desorbed into the gas chroraatograph, empty
      the purging chamber.  Wash the chamber with a minimum of two 5 mL
      flushes of reagent water to avoid carry-over of pollutant compounds.
      Baking out the purging chamber to remove  traces of possible
      contamination is highly recommended.

7.11  After desorbing the sample for 4 minutes, recondition the trap by
      returning the purge and trap device to the purge mode.  Wait 15
      seconds then close the syringe valve on the purging device to begin
      gas flow through the trap.  The trap temperature should be maintained
      at 180*C.  Trap temperatures up to 220"C may be employed, however,
      the higher temperature will shorten the useful life of the trap.
      After approximately 7 minutes, turn off the trap heater and open the

                                 VOA D-30                              Rev.  9/88

-------
      syringe valve to stop the gas flow through the trap.  When cool, the
      trap is ready for the next sample.

7.12  If the initial analysis of a sample or a dilution of a sample has
      concentration of target compounds that exceeds the initial
      calibration range, the sample must be reanalyzed at a higher
      dilution.  When a sample is analyzed that has saturated ions from a
      compound, this analysis must be followed by a blank reagent water
      analysis.  If the blank analysis is not free of interferences, the
      system must be decontaminated.  Sample analysis may not resume until
      an acceptable blank can be analyzed that is free of interferences.
      (Exhibit E, Part III)

7.13  All dilutions must be made such that major constituents are not
      saturated but give a response greater than that of the nearest
      internal standard.

7.14  Internal standard responses and retention times in all samples and
      standards must be evaluated immediately after or during data
      acquisition.  If the retention time for any internal standard changes
      by more than 30 seconds from the latest daily (12 hour) calibration
      standard, the chromatographic system must be inspected for
      malfunctions and corrections made as required.  If the extracted ion
      current profile (EICP) area for any internal standard changes by more
      than a-;factor of two (-50% to +100%), the mass spectrometric system
      must be inspected for malfunction and corrections made as
      appropriate.  When corrections are made,  reanalysis of samples
      analyzed while the system .was malfunctioning is necessary.  Internal
      Standard Areas and Retention Times are reported on Form VIII HCV, as
      described in Exhibit B.

8.     Qualitative Analysis

8.1   Criteria for Verification of Target Compound Identifications

      8.1.1   The target compounds  listed  in Exhibit C shall be identified
              by an analyst  competent  in the interpretation of mass spectra
              by comparison  of  the  sample 'mass spectrum  to the mass
              spectrum of a  standard of the suspected compound.

      8.1.2   Two criteria must be  satisfied to verify the identifications:

              8.1.2.1    Elution of the sample component at the same GC
                         relative retention time as the standard component,

              8.1.2.2    Correspondence of the sample component and
                         standard component mass spectra.

      8.1.3   For establishing  correspondence of the GC  relative retention
              time (RRT), the sample component RRT must  compare within
              ±0.06 RRT units (or ±5 seconds) of the RRT of the standard
              component.  For reference, the calibration standard must be
              run on the same shift as the sample.  If coelution of
              interfering components prohibits accurate  assignment of the

                                 VOA D-31                             Rev. 9/88

-------
I
                              sample component RRT from the total ion chroraatogram, the RRT
                              should be assigned by using extracted ion current profiles
                              for ions unique to the component of interest.

                      8.1.4   For comparison of standard and sample component mass spectra,
                              mass spectra obtained on the contractor's GC/ MS are
                              required.  Once obtained, these standard spectra may be used
                              for identification purposes, only if the contractor's GC/MS
                              meets the daily tuning requirements for BFB  (Table 2).  These
                              standard spectra may be obtained-from the daily calibration
                              standard or from the user-created mass spectral library.  The
                              requirements for qualitative verification by comparison of
                              mass spectra are as follows:

                              8.1.4.1    All ions present in the standard mass spectrum at
                                         a relative intensity greater than 10% (most
                                         abundant ion in the spectrum equals 100%) must be
                                         present in the sample spectrum.

                              8.1.4.2    The relative intensities of ions specified in the
                                         above paragraph must agree within ±20% between the
                                         standard and sample spectra.  (Example:  For an
                                         ion with an abundance of 50% in the standard
                                         spectrum, the corresponding sample abundance must
                                  •be between 30 and 70 percent).

                              8.1.4.3    Ions greater-than 10% in .the sample spectrum but
                                         not present in the standard spectrum must be
                                         considered and accounted for by the analyst making
                                         the comparison.  Both raw and background
                                         subtracted spectra must be evaluated.  The
                                         verification process should favor false positives.
                                         All compounds meeting the identification criteria
                                         must be reported with their spectra.  For all
                                         compounds below the CRQL report the actual value
                                         followed by a "J", e.gv "3J."

                      8.1.5   If a compound cannot be verified by all of the criteria in
                              8.1.4.3, but in the technical judgement of the mass spectral
                              interpretation specialist, the identification  is correct,
                              then the Contractor shall report that identification and
                              proceed with quantification  in 9.

                8.2   A library search shall be executed for non-target sample components
                      for the purpose of tentative identification.  For this purpose,  the
                      most recent available version of the NBS Mass Spectral Library shall
                      be used.  Computer generated library search routines should not use
                      normalization routines that would misrepresent the library or unknown
                      spectra when compared to each other.

                      8.2.1   Up  to  10 nonsurrogate organic compounds of greatest apparent
                              concentration not  listed  in  Exhibit C  shall  be  tentatively
                               identified via  a  forward  search  of  the  NBS Mass  Spectral
                               Library.   (Substances with  responses  less  than 10%  of the
                                                 VGA D-32
Rev. 9/88

-------
               internal  standard  are not  required  to be searched  in  this
               fashion).   Only  after visual comparison of sample  spectra
               with  the  nearest library searches will the mass spectral
               interpretation specialist  assign a  tentative  identification.

      8.2.2    Guidelines  for making tentative identification are as
               follows:

               8.2.2.1    Major ions in the reference spectrum (ions greater
                         than  10% of the most abundant ion) should be
                         present in the sample spectrum.

               8.2.2.2    The relative intensities of the major ions should
                         agree within ±20%.  {Example:   For an ion with an
                         abundance of 50 percent in the standard spectrum,
                         the corresponding sample abundance must be between
                         30 and 70 percent.)

               8.2.2.3    The molecular ion of the reference spectrum must
                         be present in sample spectrum.

               8.2.2.4    Ions present in the sample spectrum but not in the
                         reference spectrum should be reviewed for possible
                         background contamination or presence of co-eluting
                         compounds.

               8.2.2.5    Ions present in the reference spectrum but not in
                         the sample spectrum should be reviewed  for
                         possible subtraction from the sample spectrum
                         because of background contamination or co-eluting
                         compounds.   Data system library reduction programs
                         can sometimes create these discrepancies.

      8.2.3    If  in the opinion of the mass spectral specialist, no valid
               tentative identification can be made, the compound should be
               reported  as unknown.  The mass spectral specialist should
               give  additional  classification of the unknown compound, if
               possible  (i.e. unknown aromatic, unknown hydrocarbon,. unknown
               acid  type, unknown chlorinated compound).  If probable
               molecular weights can be distinguished, include them.

9.    Quantitative Analysis

9.1   Target compounds identified shall be quantified by  the internal
      standard method.   The internal standard used shall  be the one
      assigned in Table 2.1,  Volatile  Internal Standards  with Corresponding
      Target Analytes Assigned for Quantitation (Exhibit  E).  The EICP area
      of the characteristic ions  of analytes listed in Table 3  is used.
      The relative response factor (RRF)  from the daily calibration
      standard analysis is used to calculate the  concentration in the
      sample.  Use the relative response  factor as determined in paragraph
      6.3.3 and calculate  the analyte  concentration in the  sample using
      Equation D3.
                                 VGA D-33                             Rev.  9/88

-------
      Xylenes  (o,  m,  and p isomers}  are  to  be  reported as  total xylenes.
      Since o-  and p-xylene overlap,  the two xylene peaks  (o/p and m)  must
      be added together to give the  total concentration of all xylene
      isomers.

      Equation D3
      Sample Concentration (mg/kg)   -
      Where:
                                       (Ais)(RF)(Vi)(Ws)
      AJJ  -  Area of the characteristic ion for the
             compound to be measured.

      A£S -  Area of the characteristic ion for the
             specific internal standard from Table 3.

      Ig  -  Amount of internal standard added in
             nanograms (ng).

      Vfc  -  Volume of total extract (uL).

      V^  -  Volume of extract added (uL) for purging.

      Wg  -  Weight of sample extracted (mg).

9.2   An estimated concentration for non-target compounds tentatively
      identified shall be quantified by the internal standard method. For
      quantification, the nearest internal  standard free of interferences
      must be used.

      Equation D3 is used to calculate concentrations.  Total area counts
      (or peak heights) from the total ion  chromatograms are to be used for
      both the compound to be measured and  the internal standard.  A
      relative response factor (RRF) of one (1) is to be assumed.  The
      value from this quantitation shall be qualified as estimated.  This
      estimated concentration should be calculated for all tentatively
      identified compounds as well as those identified as unknowns.

9.3   Use Equation D4 to calculate surrogate standard recoveries on all
      samples, blanks and spikes.  Determine if recovery is within
      specified limits.

            Equation D4
Percent Surrogate Recovery  «-
                                                X  100%
                                 VOA D-34
                                                           Rev.  9/88

-------
            Where:

            Qd  -  Quantity determined by analysis

            Qa  -  Quantity added to sample

9.4   If recovery is not within specified limits,  the following is
      required:

      9.4.1    Check to be sure there are no errors in calculations,
               surrogate solutions, and internal standards.  Also, check
               instrument performance.

      9.4.2    Recalculate the sample data  if any of  the above checks reveal
               a problem.

      9.4.3    If  surrogates cannot be detected due to dilution factors,
               this requirement need not be met.  (Report the value as DL.)

               NOTE:  Surrogate recovery limits are MANDATORY for method
               blanks.  However, surrogate  recoveries for high concentration
               samples and control matrix spikes are ADVISORY at this time.
               Reanalysis of samples and control matrix spikes is not
               required if surrogate recoveries are outside of contractually
               specified limits.  However,  the analyst must, verify that the
               deviation is not a result of laboratory error.
                                 VOA D-35
Rev. 9/88

-------

-------
                               EXHIBIT  D
ANALYTICAL PROCEDURES FOR HIGH CONCENTRATION EXTRACTABLE ORGANIC WASTE
                                SAMPLES
                                                               Rev. 9/88

-------

-------
                               TABLE OF CONTENTS
 SECTION I  -  INTRODUCTION
   Page






EXT D-l
.SECTION II  -  SAMPLE STORAGE AND  PREPARATION   	  EXT  D-3
 SECTION  III  -  SCREENING  FOR  EXTRACTABLE  TARGET  COMPOUNDS  	 EXT D-17
 SECTION  IV  - GC/MS ANALYSIS OF EXTRACTABLE TARGET COMPOUNDS  	 EXT D-21
                                                                    Rev.  9/88

-------

-------
                                   SECTION I
                                 INTRODUCTION
The  samples received  for high concentration analysis will consist of  three
types of phases:  solids, water  immiscible liquids, and water miscible
liquids.   It  is also  possible to receive samples that will be multiple phase
such as soil, water and oil  in  the same sample jar.  Because of this
possibility,  the  samples are to be "phase separated" into their individual
phases.  The  weight of the aliquot of each phase analyzed, and the phase type
are  recorded  on appropriate data sheets and reported with the sample data.

The  phase  separation  techniques employed will vary according to the types of
samples received.  Since it is  impossible to know the number and types of
phases that will be present in a sample, the choice of phase separation
techniques is left to the discretion of the analyst.  Various techniques can
be employed to separate the phases.  These include pipetting off liquid
phases (decanting should not be done), centrifuging to remove suspended
solids, use of spatulas to remove solids (wooden tongue depressors work
well).  All operations should be done with disposable phthalate-free labware.
This eliminates the problem of cleaning contaminated glassware.  The separate
phases should be stored into glass containers with teflon-lined screw caps.
This allows for storage and handling of the waste in a safe manner.   Under no
circumstances are samples to be homogenized to eliminate separate phases:

Each individual phase is treated as a subsample,  .and is taken through the
separate GC/MS and GC/ECD procedures as a sub-sample.  The results of each of
the analyses are to be reported for each phase of the sample.

The analytical methods that follow are designed to analyze the organic
compounds on the Target Compound List (TCL) (See Exhibit C).   The methods are
divided into the following sections:   sample preparation, screening,  and
analysis.   Sample preparation covers sample storage. .
                                 EXT D-l
Rev. 9/88

-------
1.    Method For The Determination of Extractable Organic Compounds

1.1   Scope and Application

      These procedures are designed for the preparation of waste samples
      which may contain organic chemicals at a level greater than 20 mg/kg.

      1.1.1   The extracts prepared using  these methods  are screened by
              GC/FID  for  Base/Neutrals, Acids and Pesticides/PCBs  target
              compounds.  The  results of these screens will determine whether
              sufficient  quantities of pollutants are present to warrant
              analysis by the  high or low/medium protocol.
                                  EXT D-2             •               Rev.  9/88

-------
          SECTION II
SAMPLE STORAGE AND PREPARATION
         EXT D-3
Rev. 9/88

-------
PART A - SAMPLE STORAGE

 1.   Procedures for Sample Storage

1.1   The samples must be protected from light and refrigerated at 4*C (+2*C)
      from the time of receipt until extraction and analysis.

1.2   After analysis, extracts and unused sample volume must be protected
      from light and refrigerated at 4'C (±2°C) for the periods specified in
      the contract schedule.

PART B - SAMPLE PREPARATION FOR EXTRACTABLES

1.    Summary of Method

      Approximately 1.0 g portions of the single phase unit are transferred
      to vials and extracted with methylene chloride and are cleaned-up using
      GPC.  The methylene chloride extract, prepared using Gel Permeation
      Chromatography may be prepared in methylene chloride or a 1:1 mixture
      of methylene chloride/butyl chloride, at the option of the laboratory,
      prior to GPC cleanup.  Following GPC cleanup, a 1 mL aliquot of the
      collected fraction is screened for target compounds by GC/FID (See
      Section 9).  If organic compounds are detected by the high level
      screen, the methylene chloride extract is analyzed by GC/MS for the
      extractable target compounds listed in Exhibit C.  If organic compounds
      are not detected by the high level screen, contact SMO immediately.

2.    Interferences

      Method interferences may be caused-by contaminants in solvents,
      reagents, glassware, and other sample processing hardware'that .lead to
      discrete artifacts and/or elevated baselines in the total ion current
      profiles.  ALL of these materials must be routinely demonstrated to be
      free from interferences under the conditions of the analysis by running
      laboratory method blanks.  Matrix interferences may be caused by
      contaminants that are' coextracted from the sample.  The extent of
      matrix interferences will vary considerably from source to source,
      depending upon the nature and diversity of the samples.

3.    Limitations

3.1   The extraction procedure is designed to allow detection limits for
      screening purposes as low as 20 mg/kg for Base/Neutrals, Acids,
      chlorinated pesticides and individual PCB congeners.  For analysis
      purposes, the detection limits for Base/Neutrals, Acids, chlorinated
      pesticides and individual PCB congeners range from 20-200 mg/Kg (see
      Ex. C).  The pesticides and PCBs tend to have low responses when using
      a GC/FID detector.  Peaks that are present where pesticides and PCBs
      may elute must be viewed with extra caution; the concentration may be
      higher than the response would indicate.  If peaks are present in the
      GC/FID screens, high level analysis by GC/MS is required.  Some samples
      may contain high concentrations of chemicals that interfere with the
      analysis of other components at lower levels.
                                  EXT D-4
Rev. 9/88

-------
3.2   These extraction and preparation procedures were developed for rapid
      and safe handling of high concentration hazardous waste samples.  The
      design of the methods does not stress efficient recoveries or low
      limits of detection of all components.  Rather, the procedures were
      designed to screen at moderate recovery and sufficient sensitivity a
      broad spectrum of organic chemicals.  The results of the analyses thus
      may reflect only a minimum of the amount actually present in the
      sample.

4.    Reagents

4.1   Sodium Sulfate - Powdered anhydrous and reagent grade, heated at 400"C
      for four hours, cooled in a desiccator, and stored in a glass bottle.
      Baker anhydrous powder, catalog number 73898 or equivalent.

4.2   Methylene chloride - Pesticide residue analysis grade, or equivalent.

4.3   Methanol - Pesticide residue analysis grade, or equivalent.

4.4   Toluene - Pesticide residue analysis grade, or equivalent.

4.5   Benzene - Pesticide residue analysis grade, or equivalent.

4.6   Butyl chloride (1-chlorobutanol) - Analytical reagent grade, or
      equivalent (optional).

4.7   Base/Neutral and Acid Surrogate Standard Spiking Solution

      Prepare a solution in methanol containing the following compounds at a
      concentration of 1000 ug/mL for base/neutral surrogates, and 2000 ug/mL
      for the acid surrogate standards:
                     Base/Neutrals
                     nitrobenzene-d^
                     p-terpheny1-d^4
                     2 -fluorobiphenyl
phenol-dg
2,4,6-tribromophenol
2 -fluorophenol
4.8   Base/Neutral and Acid Control Matrix Spiking Solution.

      Prepare a spiking solution in methanol that contains the following
      compounds at a concentration of 1000 ug/mL for base/neutrals and 2000
      ug/mL for acids:

                    Base/Neutrals

                     1,2,4-trichlorobenzene
                     acenaphthene
                     2,4- dini tro to luene
                     pyrene
                     N-nitroso-di-n-propylamine  4-nitrophenol
                     1,4-dichlorobenzene
pentachlorophenol
phenol
2-chlorophenol
4-chloro-3-methylphenol
                                 EXT D-5
                  Rev. 9/88

-------
4.9   Pesticide/PCB Control Matrix'Spiking Solution.

      Prepare a spiking solution in methanol that contains the following
      pesticides in the concentrations specified:
                    Pesticide

                    heptachlor
                    dieldrin
ug/mL

1,000
1,000
4.10  Corn oil.

4.11  Purified solid matrix, supplied by EMSL/LV upon availability.

5.    Apparatus and Materials

5.1   Glass vials, at least 20 mL, with screw cap and teflon or aluminum foil
      liner.

5.2   Spatula - Stainless steel or teflon.

5.3   Balance capable of weighing 100 g to the nearest 0.01 g.

5.4   Vials and caps, 2 mL for GC auto sampler.

5.5   Disposable Pasteur pipets, packed with glass wool rinsed with methylene
      chloride.

5.6   15-mL.concentrator tubes.

5.7   Ultrasonic cell disruptor, Heat Systems Ultrasonics, Inc., Model W-385
      SONICATOR (475 Watt with pulsing capability, No. 200 1/2 inch tapped
      disruption horn, No. 419 1/8 inch standard tapered MICROTIP probe), or
      equivalent device with a minimum of 375 Watt output capability.  NOTE:
      In order to ensure that sufficient energy is transferred to the sample
      during extraction, the MICROTIP probe must be replaced if the tip
      begins to erode.  Erosion of the tip is evidenced by a rough surface.

5.8   Sonabox acoustic enclosure - recommended with above disruptors for
      decreasing cavitation sound.

5.9   Test tube rack.

5.10  Pyrex glass wool.

5.11  pH paper range 1-14. (Fischer Part No. A979 wide range test ribbons or
      equivalent).

6.    Sample Preparation.

6.1   Transfer approximately 1 g of each phase (record weight to the nearest
      0.1 g) of the sample to separate 20 mL vials.  Wipe the mouth of the
                                 EXT D-6
                  Rev. 9/88

-------
      vial with a tissue before weighing to remove any sample material.   Cap
      the vial before proceeding with the next sample to avoid any cross
      contamination.

      6.1.1   Prepare a  1.0 g control matrix consisting of corn oil.

      6.1.2   Prepare a  1.0 g method blank consisting of purified solid
              matrix. (May be available  through EMSL/LV) .

      6.1.3   Determine  pH of sample with pH paper.  (Add 1 mL of water to 1
              g of  soil  or water immiscible liquid and shake.  Decant water
              portion and measure pH of water with pH paper.)

6.2   Add 100 uL surrogate spiking solution to each sample,  blank and the
      control matrix.

6.3   Add 100 uL of each control matrix standard spiking solution to the
      control matrix.

6.4   Add 2.0 g of powdered anhydrous sodium sulfate  to sample.

6.5   Immediately add 10 mL of methylene chloride to  the sample.

6.6   Disrupt the sample by ultrasonic probe for 2 minutes at 100 watts
      power .

6.7   Loosely pack disposable Pasteur pipets with 2-3 cm glass wool  plugs.
      Filter the extract through the glass wool and collect  a minimum of 8.0
      »L of the extract in a concentrator tube, __

6.8   Proceed to extract clean-up by GPC, Section 7.

7.     Apparatus and Materials for Gel Permeation Chromatographv (GPC) Clean
7.1   Beakers, 400 mL.

7.2   Kuderna- Danish (K-D) apparatus.

      7.2.1   Concentrator tube - 10 mL, graduated (Kontes K-570040-1029 or
              equivalent) .

      7.2.2   Evaporative flask - 500 mL (Kontes K- 570001 -0500 or
              equivalent) .
                                •

      7.2.3   Snyder  column  - three-ball macro (Kontes K-503000-0121 or
              equivalent) .

      7.2.4   Snyder  column  - two ball micro (Kontes K- 569001 -0219 or
              equivalent) .

7.3   Silicon carbide boiling chips  - approximately 10/40 mesh.   Heat to
      400*C for 30 minutes or Soxhlet extract with methylene chloride.


                                 EXT D-7                           Rev. 9/88

-------
7.4   Water bath - heated, with concentric ring cover, capable of temperature
      control (±2"C).  The bath should be used in a hood.

7.5   Top loading balance, capable of accurately weighing 0.01 g.

7.6   Balance - Analytical, capable of accurately weighing 0.0001 g.

7.7   Nitrogen evaporation device equipped with a water bath that can be
      maintained at 35-40*C.  The N-Evap by Organomation Associates, Inc.,
      South Berlin, MA (or equivalent) is suitable.

7.8   0.45 micron teflon membrane filter, stainless steel and/or teflon
      filter holder with a 13 to 25 diameter (Millipore, Schleicher and
      Schuell or equivalent).

7.9   Syringes, 10 mL glass with luer lock tip.

7.10  Gel permeation chromatography cleanup device.   (Automated system.)

      7.10,1   Gel permeation chromatograph  (GFC)  Analytical  Biochemical  Labs,
               Inc.,  Columbia, MO,  GPC Autoprep 1002  or equivalent.

      7.10.2   25  mm ID X 700  mm glass column.

      7.10.3   Optional 5 cm guard column (Supelco 5-8319 or  equivalent)  with
               appropriate fittings to connect  to  the inlet side  of  the
               analytical column.

      7.10.4   70  g 200-400 mesh .Bio Beads (S-X3),  Bio Rad Laboratories,
               Richmond,  CA Catalog 152-2750  or equivalent (an additional 5  g
               of  Bio Beads is required  if the  optional guard column is
               employed).

      7.10.5   Fixed wavelength  ultraviolet detector  (254 nm)  with a semi-prep
               flow-through cell.

      7.1016   Strip chart recorder.

8.    Reagents

8,1   Methylene chloride, pesticide residue analysis grade,  or equivalent.

      8.1.1    Some brands of  methylene  chloride may  contain  unacceptably high
               levels  of  acid  (HC1).   Check the pH  by shaking equal  portions
               of  methylene chloride  and water,  then  check the  pH of the  water
               layer.

               If  the  pH  of the  raethylene chloride  is  less than or equal  to  5,
               filter the solvent  through a 2 in. X 15 in. glass column
               containing activated basic alumina.  This  column should be
               sufficient for  processing approximately 20-30  liters  of
               solvent.   Alternatively a different  source of  methylene
               chloride should be  found.
                                  EXT  D-8
Rev. 9/88

-------
8.2   Cyclohexane, pesticide residue analysis grade,  or equivalent.

8.3   GPC solution

      8.3.1   Calibration solution prepared  in methylene chloride containing
              the following analytes (in order of elution):

                                           mg/mL
              polystyrene   (MW 300,000)      2.8  (Optional)
              corn oil                     60.0
              bis(2-ethylhexyl)phthalate     3.0
              pentachlorophenol              0.3
              perylene                       0.02

      8.3.2   Store the calibration solution in an amber glass bottle with a
              teflon lined screw cap at 4*C,  and protect from light.
              (Refrigeration may cause the corn oil to precipitate.  Allow
              the calibration solution to stand at room temperature until the
              corn oil dissolves.)  Replace  the calibration standard solution
              as necessary or a minimum of every 3 months.

9.    Extract Cleanup

      The GC screen may be performed before the GPC column cleanup.   The
      results from the screen will indicate how much sample should be loaded
      onto the GPC column so as not to contaminate the column and,  at the
      same time, have enough extract for GC/MS analysis.   This will preserve
      the life of the GPC column since these columns are easily contaminated
      by high concentration samples.

9.1   GPC Setup and Calibration

      9.1.1   Column Preparation

              9.1.1.1    Weight 70.0 g (75.0  g if using optional guard
                         column) of Bio Beads (S-X3) into a 400 mL beaker.

              9.1.1.2    Add an excess quantity of a 1:3 (v/v mixture) of
                         methylene chloride/cyclohexane to ensure that the
                         beads will be completely submerged during the
                         swelling process.   Cover the beaker to prevent
                         solvent evaporation and allow the beads to swell for
                         a minimum of 16 hours.

              9.1.1.3    Prepare a slurry of approximately 50% swelled beads
                         and 50% solvent.   Turn the column upside down from
                         its normal position, and remove the inlet bed
                         support plunger (the inlet plunger is longer than
                         the outlet plunger).  Position and tighten the
                         outlet bed support plunger as near to the column end
                         as possible (approximately 2.5 cm from the end).
                                 EXT D-9                           Rev. 9/88

-------
9.1.1.4    Ensure that the outlet stopcock is closed.  Place a
           small amount of solvent in the column to minimize
           the formation of air bubbles at the base of poured
           column packing.   Swirl the bead/solvent slurry to
           get a homogeneous mixture, and quickly and
           consistently pour the swelled slurry into the glass
           column.   Never allow the beads to settle completely
           while pouring the column.   A glass rod can be used
           to facilitate pouring the mixture down the side of
           the column and to help minimize bubble formation.
           Use additional solvent to rinse the beaker and
           transfer all of the beads to the column.

9.1.1.5    When the transfer solvent has drained into the
           column bed, the inlet bed support plunger can be
           used to temporarily compress the gel bed.  Do not
           tighten the column/plunger seal if beads are caught
           between it and the glass column.  Remove the plunger
           and rinse it and the sides of the glass column with
           minimal solvent.   A clean paper towel can be used to
           remove any remaining beads.

9.1.1.6    Insert the inlet bed support bed plunger, compress
           the gel bed approximately 1 cm and tighten.

9.1.1.7    Pack the optional 5 cm guard column with
           approximately 2-3 g of the remaining pre-swelled
           beads (different guard columns may require different
           amounts:).  Connect the guard column to the inlet of
           the analytical column.

9.1.1.8    Connect the column inlet to the solvent reservoir
           (reservoir should be placed higher than the top of
           the column) and place the column outlet tube in a
           waste container.   Pump methylene chloride through
           the column at a rate of 5 mL/min for 1 hour.  While
           the solvent in the column is being changed from
           methylene chloride/cyclohexane to methylene chloride
           only, the gel may pack down or undergo additional
           swelling, which will change the pressure.

9.1.1.9    After washing the column for at least one hour.,
           connect the column outlet tube to the inlet side of
           the UV detector.   Connect the system outlet to the
           outlet side of the UV detector.   While still pumping
           methylene chloride through the column,  adjust the
           inlet bed support plunger until approximately 6-10
           psi backpressure  is achieved.

9.1.1.10   When the GPC column is not used for several days,
           evaporation of methylene chloride may cause  column
           drying,  which can cause stratification or channeling
           in the gel.  To prevent this drying,  connect the
           column outlet line to the  column inlet.   This will

                   EXT D-10                           Rev. 9/88

-------
                   allow a continuous recycling of the solvent when the
                   system is not in use.  If channeling occurs, the gel
                   must be removed from the column, reswelled, and
                   repoured as described above.  If drying occurs,
                   methylene chloride should be pumped through the
                   column until the observed column pressure is
                   constant'and the same as when the column was new.
9.1.2   Calibration of the GPC column

        9.1.2.1    Using a 10 mL syringe, load sample loop #1 with
                   calibration solution (8.3.1). With the ABC automated
                   system, the 5 mL sample loop requires a minimum of 8
                   mL of the calibration solution.  Use a firm,
                   continuous pressure to push the sample onto the
                   loop.

        9.1.2.2    Inject the calibration solution and obtain a UV
                   trace showing a discrete peak for each component.
                   Adjust the detector and/or recorder sensitivity to
                   produce a UV trace similar to Figure 1 that meets
                   the following requirements.   Differences between
                   manufacturer's cell volumes detector sensitivities
                   may require a dilution of the calibration solution
                   to achieve similar results.   An analytical flow-
                   through detector cell will require a much less
                   concentrated solution than the semi-prep cell, and
                   therefore the analytical cell is n.ot acceptable for
                   use.

                          UV Trace Requirements:

                         o    Peaks must be observed for all compounds
                              in the calibration solution.

                         o    Perylene peaks must not be saturated and
                              must exhibit baseline resolution.

                         o    Corn oil and phthalate peaks  must exhibit
                              baseline resolution.
        9.1.2.3    Using the information from the UV trace, establish
                   appropriate collect and dump time periods to ensure
                   collection of all target analytes.  Initiate column
                   eluate collection just before elution of bis(2-
                   ethyIhexyl)phthalate (approximately 30 minutes) and
                   after the elution of the corn oil (approximately 20
                   minutes).  Stop eluate collection shortly after the
                   elution of perylene (approximately 50 minutes).
                   Collection should be stopped before sulfur would
                   elute, if it were present (approximately 55
                   minutes).  Each laboratory is required to establish
                   its specific time sequences.  The times  provided are
                   for general guidance only.
                           EXT D-ll
Rev. 9/88

-------
                         NOTE: The collect and dump times must be adjusted to
                         compensate for the difference in volume of the lines
                         between the UV detector cell and the collection
                         flask.

              9.1.2.6    Verify the flow rate by collecting column eluate for
                         10 minutes in a graduated cylinder and measure the
                         volume, which should be 45-55 mL (4.5-5.5 mL/min).
                         If the flow rate is outside of this range,
                         corrective action must be taken, as described above.
                         Once the flow rate is within the range of 4.5-5.5
                         mL/min, record the column pressure and room
                         temperature.  Changes in pressure, solvent flow
                         rate, and temperature conditions can affect analyte
                         retention times and must be monitored.

              9.1.2.7    Reinject the calibration solution after appropriate
                         collect and dump cycles have been set and the
                         solvent flow and column pressure have been
                         established.

                         9.1.2.7.1    Measure and record the volume of
                                      collected GPC eluate in a graduated
                                      cylinder.   The volume of GPC eluate
                                      collected for each sample extract
                                      processed may be used to indicate
                                      problems with the system during sample
                                      processing.

                         9.1.2.7.2    The retention times for bis(2-
                                      ethylhexyl)phthalate and perylene must
                                      not vary more than plus or minus 5%
                                      between calibrations.   If the retention
                                      time shift is greater than 5%,  take
                                      corrective action.

9.2   GPC Extract Cleanup

      9.2.1   Studies have demonstrated that the recovery of certain aromatic
              compounds  is improved when samples are introduced into the GPC
              in a 1:1 solvent mixture of methylene chloride/butyl chloride.
              If it is used, reduce the 8.0 mL sample volume to 3.5-4.0 mL
              under a stream of dry nitrogen (Paragraph 8.4.3).  Reconstitute
              to the 8.0 mL volume with butyl chloride.  Thoroughly mix the
              sample before proceeding.

      9.2.2   Pre-filter each sample extract through a 0.45 micron PTFE
              filter (Millipore, Schleicher and Schuell or equivalent).
              Samples that contain suspended particles must be centrifuged
              prior to filtration.  Clean the filter holder assembly between
              samples.
                                 EXT D-12
Rev. 9/88

-------
9.2.3   Using a  10 mL syringe,  load the nominal 5 mL sample loop.

        CAUTION:  Approximately 2 mL of the extract remains in the
        lines between the  injection port and the sample loop; excess
        sample also passes  through the sample loop to waste.

        Attach the syringe  to the turn lock on the injection port.  Use
        firm, continuous pressure to push the sample onto  the loop.  If
        the  sample is difficult to load, some part of the  system may be
        blocked.  Take appropriate corrective action.

9.2.4   After loading a loop, and before removing the syringe from  the
        injection port index the GPC to the next loop.  This will
        prevent  loss of sample  caused by unequal pressure  in the loops,

9.2.5   After loading each  sample loop, wash the loading port with
        methylene chloride  in a PTFE wash bottle to minimize cross
        contamination.  Inject  approximately 10 mL of methylene
        chloride to rinse  the common tubes.

9.2.6   Column overloading.

        9.2.6.1    Column overloading can occur when too much material
                   is loaded in a sample loop.

                   9.2.6.1.1    For highly contaminated samples,  dilute
                                the extract and'process  in more  than
                                one sample loop.   An example dilution
                                procedure is to mix  10 mL of sample
                                extract with 10 mL of methylene
                                chloride or 1:1 butyl
                                chloride/methylene chloride,  shake  well.
                                to thoroughly mix, and load into  two
                                sample  loops.

                   9.2.6.1.2    After GPC cleanup,  combine the
                                collected fractions  and treat as  a
                                single  sample.   Therefore,  no
                                additional dilution  factor will be
                                required when a sample extract is
                                diluted and divided  for  GPC cleanup.

9.2.7   After loading all the sample loops, index the GPC  to the 00
        position, switch to the "run" mode and start the automated
        sequence.

9.2.8   Process each sample using the collect and dump cycle times
        established in 9.1.2.   Process calibration standards, control
        matrix (GPC blank), samples and method blanks in the following
        order:
                           EXT D-13                           Rev.  9/88

-------
                     Loop 1    Calibration Standard
                     Loop 2    Control Matrix (corn oil)
                     Loop 3-22 Samples
                     Loop 23   Calibration Standard (Verification)

              NOTE:  GPC cleanup sequence must end with a calibration
              standard verification regardless of the number of samples
              processed.

      9.2.9   Collect each  sample  in a 250 mL Erlenmeyer flask, covered with
              aluminum foil to reduce solvent evaporation.  Monitor  sample
              volumes collected.   Do not concentrate the processed GPC
              extract at this point.  Changes in sample volumes collected may
              indicate one  or more of the following problems:

              9.2.9.1    Change in solvent flow rate, caused by channeling in
                         the column or changes in column pressure.

              9.2.9.2    Increase  in column operating pressure due to the
                         absorbtion of particles or gel fines onto either the
                         guard column or the analytical column gel,  if a
                         guard column is not used.

              9.2.9.3    Leaks in  the system or significant variances in room
                         temperature.

      9.2.10  Evaluation of calibration standards.

              9.2.10.1   Evaluate  the retention times for bis(2-         —-—^
                         ethylhexyl)phthalate and perylene in each daily
                         calibration standard UV trace.   The retention time
                         for either compound cannot exceed a 5% retention  .
                         time shift when compared to the retention time
                         established in the initial calibration (loop 1).

              9.2.10.2   Corrective action must be taken before sample
                         cleanup can proceed.   All samples processed prior to
                         an unacceptable calibration verification must be
                         identified in the case narrative.

              9.2.10.3   Acceptable GPC performance is demonstrated by the
                         successful analysis of the calibration standard
                         solution.   Corrective actions may include those
                         listed in 9.1.1 and 9.1.2 above or may require
                         repacking and recalibration of the column.

9.3   Transfer a 1.0 mL aliquot of the processed extract to a GC vial.   Do
      not discard the remaining extract until  the GC/FID screen has  been
      performed.  Proceed to GC/FID screen,  Section III.

      9.3.1   For the control matrix sample and method blank, concentrate the
              extract (9.4).
                                 EXT D-14
Rev. 9/88

-------
      9.3.2    Store all extracts at 4°C in the dark  in teflon-sealed
               containers until all analyses are performed.

9.4   If no sample peaks are detected or all are less than 10% full scale
      deflection during the GC/FID screen, the sample oust be concentrated.
      The control matrix samples and the method blanks are also concentrated.

      9.4.1    Transfer the extract (9.2.9) to a Kuderna Danish  (K-D)
               concentrator consisting of  a 10 mL concentrator tube and a 500
               mL evaporative flask.  Other concentration devices or
               techniques may be used if equivalency  is demonstrated for all
               extractable and pesticide compounds  listed in Exhibit C.

      9.4.2    Add one or two clean boiling chips to  the evaporative flask and
               attach a three ball Snyder  column.   Pre-wet  the Snyder column
               by adding about 1 mL methylene chloride to the top.  Place the
               K-D apparatus on a hot water bath (80° to 90*C) so that the
               concentrator tube is partially immersed in the hot water and
               the entire lower rounded surface of  the flask is bathed with
               hot vapor.  Adjust the vertical position of  the apparatus and
               the water temperature as required to complete the concentration
               in 10 to 15 minutes.  At the proper  rate of  distillation, the
               balls of the column will actively chatter but the chambers will
               not flood with condensed solvent.  When the  apparent volume of
               liquid reaches 1 mL, remove the K-D  apparatus and allow it to
               drain and cool for at least 10 minutes, and  make  up to 10 mL
               volume with methylene chloride.

      9.4.3    Nitrogen blowdown technique (taken from ASTM Method D 3086).

               The following method must be used for  final  concentration of
               the extracts.  Place the concentrator  tube in a warm water bath
               (35*C) and evaporate the solvent volume to just below 1 mL
               using a gentle stream of clean, dry  nitrogen (filtered through
               a column of activated carbon).

               CAUTION:  New plastic tubing must not  be used between the
               carbon trap and the sample, since it may introduce
               interferences.  The internal wall of the tube must be rinsed
               down several times with methylene chloride during the
               operation, and the final volume brought to 1.0 mL with
               methylene chloride.  During evaporation, the tube solvent level
               must be just above the water level of  the bath.  The extract
               must never be allowed to become dry.

9.5   Store all extracts at 4°C in the dark in teflon-sealed containers until
      all analyses are performed.
                                 EXT D-15                           Rev.  9/88

-------
        "E
        CO
        •o
        C
        CO
        4ml
        CO

        C
        o
        co
        o
        o
        Q.
        O
                       6ui
6ui
                    6iu OOC   HO
                       Bui fr|.  OOO'OOC MW
                     I      i       I     I      I      I
                     co
                                             to
                                                                o
                                                                to
                                                                in
                                                                in
                                                                o
                                                                in
                                                  V)
                                                  o
                                              in  3
                                              «  C
                                                                o
                                                                CO
                                              in
                                                                O
                                                                CM
                                                                in
                  s)|OAjH!iii  'esuodsay
Figure 1.  UV Detector Trace Produced by GPC Calibration Solution.


                             EXT D-16                        Rev.  9/88

-------
                   SECTION III
GC/FID SCREENING FOR EXTRACTABLE TARGET COMPOUNDS
                  EXT D-17
Rev. 9/88

-------
1.    Summary of Method

      The solvent extracts of the single phase units are screened on a gas
      chromatograph/flame ionization detector (GC/FID) using a fused silica
      capillary column (FSCC).   The results of the screen will determine the
      volume of extract taken for GC/MS analysis.

2.    Apparatus and Materials

2.1   Gas chromatograph - An analytical system complete with a temperature
      programmable gas chromatograph and all required accessories including
      syringes, analytical columns, and gases.  The injection port must be
      designed for splitless injection using capillary columns.

      2.1.1   Above GC equipped with a flame  ionization detector.

      2.1.2   GC column  - 30 m X 0.32 mm,  1 micron film thickness, silicone
              coated, fused silica capillary  column (J & W Scientific DB-5 or
              equivalent).

3.    Reagents

3.1   Methylene chloride - pesticide residue analysis  grade or equivalent.

3.2   GC calibration standard.   Prepare a stock standard solution containing
      phenol, phenanthrene,  and di-n-octylphthalate in methylene chloride.

      3.2.1   Stock standard solutions (1.00  ug/uL) - Stock standard
              •solutions can be prepared from  pure standard materials or
              purchased solutions.

              3.2.1.1    Prepare stock standard solutions by accurately
                         weighing about 0.0100 g of pure material.  Dissolve
                         the material in pesticide quality methylene chloride
                         and dilute to volume in a 10 mL volumetric flask.
                         Larger volumes may be used at the convenience of the
                         analyst.  If compound purity is assayed at 96% or
                         greater, the weight may be used without correction
                         to calculate the concentration of the stock
                         standard.   Commercially prepared stock standards may
                         be used at any concentration if they are certified
                         by the manufacturer or by an independent source.

              3.2.1.2    Transfer the stock standard solutions into teflon-
                         sealed screw-cap bottles.  Stored standard solutions
                         should be checked frequently for signs of
                         degradation or evaporation,  especially just prior to
                         preparing calibration standards from them.  Stock
                         standard solutions must be replaced after six months
                         or sooner if comparison with quality control check
                         samples indicates a problem.
                                 EXT D-18                          Rev. 9/88

-------
              3.2.1.3    Prepare a working standard of the GC calibration
                         standard in methylene chloride.  The concentration
                         must be such that the volume injected equals 50 ng
                         of each compound.  The storage and stability
                         requirements are the same as specified in 3.2.1.2.

4.    GC Calibration

4.1   At the beginning of each 12 hour shift,  inject the GC calibration
      standard.   The following criteria must be met:

      4.1.1   Standardized for 50% full scale response for 50 ng of
              phenanthrene.

      4.1.2   Adequately separates phenol from the  solvent front.

      4.1.3   Minimum of 25% full scale response for 50 ng of di-n-
              octylphthalate.

5.    GC/FID Screening

5.1   Suggested GC operating conditions:

      Initial Column Temperature Hold - 50*C for 4 minutes.

      Column Temperature Program - 50° to 280°C at 8 degrees  per minute.

      Final Column Temperature Hold -280'C for 8 minutes.

      Injector - Grob-type;  splitless.

      Sample Volume - 1 uL to 2 uL.

      Carrier Gas - Helium at 30 cm per sec.

5.2   Inject the GC calibration standard and ensure the criteria specified in
      4.1 are met before injecting samples.   Estimate the response for 10 ng
      of phenanthrene.

5.3   Inject the extracts of all single phase units to be screened, including
      blanks (Section II, paragraphs 9.3 and 9.4).

6.    Interpretation of Chromatograros

6.1   If no sample peaks from the extract are detected or all are less than
      10% full scale deflection,  the sample must be concentrated as per
      Section II (9.4), rescreened,  and analyzed by GC/MS.  If no peaks are
      detected or all are less than 10% full scale deflection after
      concentration, contact the Sample Management Office for instructions.

6.2   If peaks are detected at greater than 10% deflection and less than or
      equal to 100% full scale deflection, proceed with GC/MS analysis of
      this extract with appropriate dilution if necessary.


                                 EXT D-19                          Rev.  9/88

-------
6.3   If peaks are detected at greater than 100% full scale deflection,
      calculate the dilution necessary to reduce the major peaks to between
      50% and 100% full scale deflection.  Use this dilution factor to dilute
      the extract for GC/MS analysis.
                                 EXT D-20
Rev. 9/88

-------
                  SECTION IV
GC/MS ANALYSIS OF EXTRACTABLE TARGET COMPOUNDS
                 EXT D-21
Rev. 9/88

-------
1.    Summary of Method

      This method is to be used for the GC/MS analysis of extractable
      extracts screened by Section III protocols.

2.    Apparatus and Materials

2.1   Gas chromatograph/mass spectrometer system.

      2.1.1   Gas chromatograph  - An analytical  system complete with a
              temperature programmable gas chroraatograph suitable  for
              splitless injection and all required accessories including
              syringes, analytical columns,  and  gases.

      2.1.2   Column  - 30 m X 0.25 mm ID (or 0.32 mm) bonded-phase silicone
              coated  fused  silica capillary  column (J & W Scientific DB-5  or
              equivalent).  A film thickness of  1.0 micron  is recommended
              because of its larger capacity.  A film thickness of 0.25
              micron  may be used.

      2.1.3   Mass  Spectrometer  - Capable of scanning from  35 to 510 amu
              every 1 second or  less, utilizing  70 volts (nominal)  electron
              energy  in the electron impact  ionization mode and producing  a
              mass  spectrum which meets all  required criteria when 50 ng of
              decafluorotriphenylphosphine (DFTPP) is injected through the GC
              inlet (Exhibit E,  Table E-l).                   .  -

              NOTE:   DFTPP  criteria must be  met  before any  sample  extracts
              are analyzed.  Any samples analyzed when DFTPP criteria have
              not been met  will  require reanalysis at no additional cost to
              the Government.

      2.1.4   Data  system - A computer system must be interfaced to the mass
              spectrometer  that  allows the continuous acquisition  and storage
              on machine readable media of all mass spectra obtained
              throughout the duration of the chromatographic program.  The
              computer must have software that allows searching any GC/MS
              data  file for ions of a specific mass and plotting such ion
              abundances versus  time or scan number.  This type of  plot is
              defined as an Extracted Ion Current Profile (EICP).   Software
              must  also be  available that allows integrating the abundance in
              any EICP between specified time or scan number limits.

3.    Reagents

3.1   Internal standards -  1,4 dichlorobenzene-d^,  napthalene-dg,
      acenaphthene-d^Q,  phenanthrene-d^Q,  chrysene-dio, perylene-d^•  An
      internal standard solution can  be prepared by dissolving 200  rag of each
      compound in 50 mL of methylene  chloride.   It  may be necessary to use  5%
      to 10% benzene or toluene  in this  solution and  a few minutes  of
      ultrasonic mixing in order to dissolve  all the  constituents.   The
      resulting solution will contain each  standard at a concentration of
                                 EXT D-22                          Rev. 9/88

-------
      4000 ng/uL.  A 10 uL portion of this solution should be added to each 1
      mL of sample extract.  This will give a. concentratidn of 40 ng/uL of
      each constituent.

3.2   Target Compound Calibration Standards

      3.2.1    Prepare  calibration standards at a minimum of three
               concentration  levels.  Each calibration standard shall contain
               each compound  of  interest and each surrogate standard.  See
               GC/MS calibration in Exhibit E  (Section 2.1.1) for calibration
               standard concentration.

4.    Calibration

4.1   Great care must be taken to maintain the integrity of all standard
     .solutions.  It is recommended that all standard solutions be stored at
      4°C or less in screw cap amber bottles with teflon liners.   Fresh
      standards should be prepared every six months at a minimum.

4.2   Each GC/MS system must be hardware tuned to meet the criteria listed in
      Table E-l in Exhibit E for a 50 ng injection of
      decafluorotriphenylphosphine (DFTPP).   No sample analyses can begin
      until all these criteria are met.   This criteria must be demonstrated
      each 12 hour shift.  DFTPP has to be injected to meet this criterion.
      Post-acquisition manipulation of abundances is not acceptable.

4.3   The internal standards selected in paragraph 3.1 should permit most
      components of interest in a chromatogram to have retention times of
      0.80 to 1.20 relative to the internal standards (see instructions for
      Form VI, Initial Calibration Data).   Use the base peak ion from the
      specific internal standard as the primary ion for quantification,  found
      in Table D-5.   If interferences are noted,  use the next most intense
      ion as the secondary ion,  i.e.  For 1,4-dichlorobenzene-d^ use ra/z 115
      for quantification.

      4.3.1    The internal standards are added to  all sample extracts just
               prior to analysis by GC/MS and  to all calibration standards.  A
               10 uL aliquot  of  the internal standard solution should be added
               to a 1 mL  aliquot of calibration standards.

4.4   Target Compound Relative Response Factors

      4.4.1    Analyze  1  to 2 uL of each calibration standard and tabulate the
               area of  the primary characteristic ion (Table D-l) against
               concentration  for each compound including the surrogate
               compounds.  Calculate relative response factors (RRF) for each
               compound us ing Equation 1.
                                 EXT D-23                   '        Rev. 9/88

-------
                     Equation 1.
                                     RRF -
                                   X  Cis
                                             is
              Where:
              AX  - Area of the characteristic  ion for the  compound to
                    be measured.

              A^g - Area of the characteristic  ion for the  specific
                    internal standard from Exhibit E.

              cis " Concentrat*-on °f tne internal standard  (ng/uL).

                    Concentr,
                    (ng/uL).
GX  - Concentration of the compound to be measured
      4.4.2   The average relative response factor  (RRF) should be calculated
              for all compounds.  A system performance check must be made
              before this calibration curve is used.  Four system performance
              check compounds  (SFCCs) are checked for a minimum average
              response factor.  These compounds are N-nitroso-di-n
              propylamine, hexachlorocyclopentadiene, 2,4-dinitrophenol, 4-
              nitrophenol.  The minimum acceptable average relative response
              factor for extractable System Performance Check Compounds is
              0.050.

      4.4.3   A maximum % Relative Standard Deviation (% RSD) of 30.0% for
              the thirteen Calibration Check Compounds (CCC) must be met for
              the calibration  curve to be valid (see Exhibit E, Section III,
              2.3.2).

      4.4.4   A check of the calibration curve must be performed once every
              12 hours during  analysis.  These criteria are described in
              detail in the instructions for Form VII HCE.  The minimum
              response factor  for the system performance check compounds is
              0.050.  If this  criteria is met, the response factor 'of all
              compounds is calculated.  A percent difference of the daily
              response factor  (12 hour) compared to the average response
              factor from the  initial curve is calculated.  A maximum percent
              difference of 25.0% is allowed for each compound flagged as
              "CCC" on Form VII.  Only after both these criteria are met can
              sample analysis  begin.

4.5   Internal standard responses and retention times in all samples must be
      evaluated immediately after or during data acquisition.   If the
      retention time for any internal standard changes by more than 30
      seconds, the chromatographic system must be inspected for malfunction
      and corrections made as required.   If the extracted ion current profile
      (EICP) area for any internal standard changes by more than a factor of
      two (-50% to +100%),  from the  latest daily (12 hour)  calibration
      standard,  the mass spectrometric system must be inspected for
      malfunction and corrections made as appropriate.   When corrections  are
      made, re-analysis of samples analyzed while the system was
      malfunctioning is necessary.
                                 EXT D-24
                                                     Rev. 9/88

-------
      Retention times and EICF areas .are reported on Form VIII.

5.    GC/MS Analysis

5.1   The following instrumental parameters are required for all performance
      tests and for all sample analyses:

               Electron  Energy     -  70  volts  (nominal)
               Mass Range          -  35  to  510  amu
               Scan Time          -  1 second per  scan

5.2   Internal standard solution is added to each sample extract.  Add 10 uL
      of internal standard solution to 1.0 mL (accurately measured) of sample
      extract.

      NOTE:  Make appropriate extract dilutions as indicated by the screening
      procedure prior to the addition of internal standards.   If any further
      dilutions of the extracts are made,  additional internal standards must
      be added to maintain the required 40 ng/uL of each constituent in the
      extract volume.   If any compound saturates the detector,  the extract
      must be diluted and reanalyzed.

      Analyze an aliquot of the 1.0 mL extract by GC/MS using a bonded phase
      silicone coated fused silica capillary column.  The recommended GC
      operating conditions to be used are  as  follows:

            Initial Column Temperature  Hold - 30*C  for 4 minutes

            Column Temperature Program      - 30* to 300°C at 8 degrees per
                                               minute

            Final Column Temperature Hold    - 300°C  for 10 minutes

            Injector Temperature            - 250°  to 300°C

            Transfer Line Temperature       - 250*  to 300*C
            Source Temperature              - According to manufacturer's
                                               specifications

            Injector                        - Grob  Type, Splitless

            Injection Volume                - 1  to  2 uL

            Carrier Gas                     - Helium at 30 cm /min

6.    •Qualitative Analysis

6.1   Single Component Extractable Target  Compounds

      6.1.1    The single component  target compounds, listed in Exhibit C,
               shall be  identified by comparison  of  the sample mass spectrum
               to the mass spectrum  of a standard of  the suspected compound.
               Two criteria must be  satisfied to verify the identifications:
               (1) elution of the sample component at the same GC relative
               retention time as the standard component, and (2)
               correspondence of the sample component and standard component
               mass spectra.

                                 EXT D-25                           Rev. 9/88

-------
      6.1.2   For establishing correspondence of the GO relative retention
              time (RRT), the sample component RRT must compare within +0.05
              of the RRT of the standard component.  For reference, the
              standard must be run on the same shift as the sample.  The RRT
              should be assigned by using extracted ion current profiles for
              ions unique to the component of interest.

      6.1.3   For comparison of standard and sample component mass spectra,
              mass spectra obtained on the Contractor's GC/MS are required.
              Once obtained, these standard spectra may be used for
              identification purposes, only if the Contractor's GC/MS meets
              the DFTPP daily tuning requirements.  These standard spectra
              may be obtained from the run used to obtain reference RRTs.

              6.1.3.1    The requirements for qualitative verification by
                         comparison of mass spectra are as follows:

                               (1)  All ions present in the  standard mass
                                    spectra at a relative intensity greater
                                    than 10% (most abundant  ion in the
                                    spectrum equals 100%)  must be  present in
                                    the sample spectrum.

                               (2)  The relative intensities  of ions
                                    specified in (1)  must agree within +20%
                                    between the standard and  sample spectra.

                                    (Example:   For an ion with an  abundance
                                    of 50%  in the  standard spectra,  the
                                    corresponding  sample  ion  abundance must
                                    be between 30  and 70%.)

                               (3)  Ions  greater than 10%  in  the sample
                                    spectrum but not present  in the  standard
                                    spectrum must  be considered and accounted
                                    for by  the analyst making the  comparison.
                                    The verification process  should favor
                                    false positives.

6.2   PCB Congener Analysis

      6.2.1   PCB's are identified and measured by level of chlorination.
              Concentration is measured for each PCB isomer group.

              The ten individual PCB congeners listed in Table D-2 are used
              as concentration calibration compounds for PCB determinations.
              One isomer at each level of chlorination is used as the
              concentration calibration standard for all isomers at that
              level of chlorination.
                                 EXT D-26
Rev. 9/88

-------
6.2.2   Identification and measurement

        Special software can be used for automated identification and
        measurement of PCBs.  Unprocessed GC/MS data are handled
        without human interaction with the software operationg on the
        dedicated computer.  A concentration for each PCB isoraer group
        is calculated automatically.

        Examine each PCB candidate spectrum after background correction
        routines have been applied.  Verify the absence of any ions
        with mass greater than the highest mass possible for the
        compound of concern.  (Ions in PCB M+ ion clusters are shown in
        Table D-3).

        6.2.2.1    For all PCB candidates,  confirm the presence of an
                   (M+70)+ ion cluster by examining ICPs or spectra for
                   at least one of the most intense ions in the
                   appropriate ion cluster.

        6.2.2.2  •  For Clo-Cly isomer groups,  examine the extracted ion
                   current profiles (EICPs) or spectra for intense
                   (M+70)+ ions that would indicate a coeluting PCB
                   containing two additional chlorines.

        6.2.2.3    For C^-Clg-PCB candidates,  examine ICPs or spectra
                   for intense (M+35)+ ions that would indicate a
                   coeluting PCB containing one additional chlorine.
                   This coelution causes interferences because of the
                   natural abundance of   C.   (This interference will
                   be small and can be neglected except when measuring
                   the area of a small amount  of a PCB coeluting with a
                   large amount of another PCB containing one more
                   chlorine).

        6.2.2.4    Use ICP data to calculate the ratio of the measured
                   peak areas  of the quantitation ion and confirmation
                   ion(s),  and compare to the  acceptable ratio (Table
                   0-3).   If acceptable ratios  are not obtained,  a
                   coeluting or partially coeluting compound may be
                   interfering.

        6.2.2.5    Quantitation and confirmation ions for each PCB
                   isomer group must maximize within +1  scan of each
                   other.
                                                                  •

        6.2.2.6    The integrated ion current  for each quantitation and
                   confirmation ion must be at  least three times
                   background  noise and must not have saturated the
                   detector.

        6.2.2.7    For each PCB isoraer group candidate,  the ratio of
                   the quantitation ion area to the confirmation ion
                           EXT D-27                           Rev. 9/88

-------
                         area must be within limits shown in Table D-3; at
                         least one ion in the (M-70)+ ion cluster must be
                         present.

6.3   A library search shall be executed for non-target compound sample
      components for the purpose of tentative identification.  For this
      purpose, the most recent available version of the EPA/NBS Mass Spectral
      Library should be used.

      6.3.1   Up  to 20 nonsurrogate organic compounds of greatest apparent
              concentration  not  listed in Exhibit C for the  extractable
              fraction shall be  tentatively identified via a forward  search
              of  the EPA/NBS Mass Spectral Library.  (Substances with
              responses  less than 10% of the nearest internal  standard are
              not required to be searched in this fashion).  Only after
              visual comparison  of sample spectra with the nearest  library
              searches will  the mass spectral interpretation specialist
              assign a tentative identification.  Computer generated  library
              search routines must not use normalization routines that would
              misrepresent the library or unknown spectra when compared to
              each other.

            6.3.2    Guidelines for making tentative identification:

                     (1)  Relative intensities  of major ioris in the reference
                          spectrum (ions greater than 10%. "of the most
                          abundant) should be present in the sample spectrum.

                     (2)  The relative intensities of the major ions  should
                          agree  within + 20%.   (Example:  For  an ion  with an
                          abundance of 50% in the standard spectra, the
                          corresponding sample  ion abundance must be  between
                          30 and 70%.

                     (3)  Molecular ions present in reference  spectrum should
                          be present in sample  spectrum.

                     (4)  Ions present in the sample spectrum  but not  in the
                          reference spectrum should be reviewed for possible
                          background contamination or presence of coeluting
                          compounds.

                     (5)  Ions present in the reference spectrum but not in
                          the sample spectrum should be reviewed for possible
                          subtraction from the  sample spectrum because of
                          background contamination or coeluting compounds.
                          Data system library reduction programs can
                          sometimes create these discrepancies.

      6.3.3   If  in the  opinion of the mass spectral specialist, no valid
              tentative  identification can be made, the compound should be
              reported as unknown.  The mass spectral specialist should give
              additional classification of the  unknown compound, if possible
              (i.e., unknown phthalate, unknown hydrocarbon, unknown  acid

                                 EXT  D-28                           Rev. 9/88

-------
               type,  unknown chlorinated compound).   If probable  molecular
               weights can be distinguished,  include them.

7 .    Quant 1 tat ion

7.1   Target Compound Quantitation

      7.1.1    The target compounds  components  identified shall be  quantified
               by the internal standard method.   The internal  standard used
               shall  be the one assigned in Table E-3,  in Exhibit E.   The EICP
               area of characteristic ions of analytes  listed  in  Tables D-l,
               D-3, and D-5 are used.  The relative  response factor (RRF) from
               the daily standard analysis is used to calculate the
               concentration in the  sample.   Secondary  ions  may be  used if
               interferences are present.   The  area  of  a secondary  ion cannot
               be substituted for the area of a primary ion  unless  a response
               factor is calculated  using the secondary ion.

               7.1.1.1    Calculate  the concentration in the sample using the
                          relative response  factor (RRF)  as  determined in
                          paragraph  4.4 and  the  following equation:
            Concentration  mg/kg =  - S - S - £
                                     (Ais)  (RRF)  (Vi) (Ws) (1000)

                         Where :

                         Vt  =•      Volume of total extract (uL)

                         V-  =      Volume of extract injected (uL)

                         Wg  =      Weight of sample extracted (grams)

                         Ax  =      Area of the characteristic ion for
                                   the compound  to be measured

                         A£S -      Area of the characteristic ion for
                                   the internal  standard

                         Ig  =      Amount of internal standard
                                   injected (ng)

                         NOTE:  Special software. may be used for automated
                         identification and measurement of PCBs.  Unprocessed
                         GC/MS data are handled without human interaction
                         with the software operating on the dedicated
                         computer.  A concentration for each PCB isomer group
                         is calculated automatically.

7.2   An estimated concentration for non- target compounds components
      tentatively identified shall be quantified by the internal standard
      method.  For quantification, the nearest internal standard free-of-
      interferences shall be used.


                                 EXT D-29                          Rev.  4/89

-------
7.3
7.2.1   The formula for calculating concentrations is the same as  in
        paragraph 7.1.1.1.  Total area counts from the total  ion
        chromatograms are to be used for both the compound  to be
        measured and the internal standard.  A relative response factor
        (RRF) of one (1) is to be assumed.  The value from  this
        quantitation shall be qualified as estimated.  This estimated
        concentration should be calculated for all tentatively
        identified compounds as well as those identified as unknowns.

Calculate surrogate standard recovery on all samples, control matrix
spikes, and method blanks.   Determine if recovery is within the
suggested limits and report on appropriate form.

7.3.1   If recovery is not within the suggested limits for  a  sample,
        the following is required:

        7.3.1.1    Check to be sure there are no errors in
                   calculations, surrogate solutions and internal
                   standards.  Also, check instrument performance.

        7.3.1.2    Recalculate the data and/or reanalyze the  extract if
                   any of the above checks reveal a problem.

        7.3.1.3    If surrogates cannot be detected due to  dilution
                   factors, the requirement in 7.3.1.2 need not be met.

7.3.2   Method Blank Surrogate Recoveries

        If one or more surrogates are outside the contract  required
        limits  (listed in Table E-5), the  laboratory must take the
        following actions:

        7.3.2.1    Check to be sure there are no errors in
                   calculations, surrogate solutions and internal
                   standards.  Also, check instrument performance.

        7.3.2.2    Recalculate the data and/or reanalyze the  extract if
                   any of the above checks reveal a problem.

        7.3.2.3    If the above measures fail to correct the  problem,
                   the analytical system must be considered out of
                   control.  The method blank and all associated single
                   phase units, including control matrix spikes must be
                   re-extracted and reanalyzed at no additional cost to
                   the agency.
                                 EXT D-30
                                                              Rev.  9/88

-------
TABLE  D-l.      Characteristic  Ions  for  Extractable Target Compounds

Parameter                                    Primary  Ion         Secondary  Ion(s)
 Phenol
 bis(-2-Chloroethyl)ether
 2-Chlorophenol
 1,3-Dichlorobenzene
 1,4-Dichlorobenzene
 Benzyl Alcohol
 1,2-Dichlorobenzene
 2-Methylphenol
 bis(2-chloroisopropyl)ether
 4-Methylphenol
 N-Nitroso-di-propylamine
 Hexachloroethane
 Nitrobenzene
 Isophorone
 2-Nitrophenol
 2,4-DimethyIpheno1
 Benzole Acid
 bis(-2-Chloroethoxy)methane
 2,4-Dichlo ropheno1
 1,2,4-Trichlorobenzene
 Naphthalene
 4-Chloroaniline
 Hexachlorobutadiene
 4-Chloro-3-methylphenol    *.
 2-Methylnaphthalene
 Hexachlorocyclopentadiene
 2,4,6-Trichlorophenol
 2,4,5-Trichlorophenol
 2-Chloronaphthalene
 2-Nitroaniline
 Dimethyl phthalate
Acenaphthylene
 3-Nitroaniline
Acenaphthene
 2,4-Dinitrophenol
 4-Nitrophenol
 Dibenzofuran
 2,4-Dinitrotoluene
 2,6-Dinitrotoluene
Diethylphthalate
4-Chlorophenyl-phenyle ther
 Fluorene
 4-Nitroaniline
 4,6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine
4-Bromophenyl-phenylether
Hexachlorob enz ene
 94
 93
128
146
146
108
146
108
 45
108
 70
117
 77
 82
139
107
122
 93
162
180
128
127
225
107
142
237
196
196
162
 65
163
152
138
153
184
109
168
165
165
149
204
166
138
198
169
248
284
   65, 66
   63, 95
   64,  130
  148, 113
  148, 113
   79, 77
  148, 113
     107
   77, 79
     107
42, 101,  130
  201, 199
   123, 65
   95,  138
   65,  109
  121, 122
   105, 77
   95,  123
   164, 98
  182, 145
  129, 127
     129
  223, 227
  144, 142
     141
  235, 272
  198, 200
  198, 200
  164, 127
  92,  138
  194, 164
  151, 153
  108, 92
  152, 154
  63,  154
  139, 65
     139
  63,  182
  89,  121
  177, 150
  206, 141
  165, 167
  92,  108
  182, 77
  168, 167
  250, 141
  142, 249
                                 EXT D-31
               Rev. 1/89

-------
TABLE D-l. (Continued)

Parameter

Pentachlorophenol
Phenanthrene
Anthracene
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3'-Dichlorobenzidine
Benzo(a)anthracene
bis(2-Ethylhexyl)phthalate
Chrysene
Di-n-octyl phthalate
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo(a)pyrene
Indeno(1,2,3-cd)pyrene
Dibenz(a, h)anthracene
Benzo(g, h, i)perylene
Alpha-BHC
Beta-BHC
Delta-BHC
Gamma-BHC (Lindane)
Heptachlor
Aldrin
Heptachlor Epoxide
Endosulfan I
Dleldrin
4,4'-DDE
Endrin
Endosulfan II
4,4'-ODD
Endosulfan Sulfate
4,4'-DDT
Methoxychlor
Chlordane-alpha
Chlordane-gamma
Endrin Retone
Primary Ion

        266
        178
        178
        149
        202
        202
        149
        252
        228
        149
        228
        149
        252
        252
        252
        276
        278
        276
        183
        181
        183
        183
        100
         66
        353
        195
         79
        246
        263
        337
        235
        272
        235
        227
        373
        373
        317
Secondary I on (...si

    264, 268
    179, 176
    179, 176
    150, 104
    101, 100
    101, 100
    91,  206
    254, 126
    229, 226
    167, 279
    226, 229

    253, 125
    253, 125
    253, 125
    138, 227
    139, 279
    138, 277
    181, 109
    183, 109
    181, 109
    181, 109
    272, 274
    263, 220
    355, 351
    339, 341
    263, 279
    248, 176
     82, 81
    339, 341
    237, 165
    387, 422
    237, 165
       228  .
    375, 377
    375, 377
    67,  319
                                 EXT D-32
                        Rev.  9/88

-------
TABLE D-2.      PCB  Congeners Used  as  Calibration  Standards
PCB Isomer Group

Concentration Calibration  Standard

Monochlorobiphenyl

Dichlorobiphenyl

Trichlorobiphenyl

Tetrachlorobiphenyl

Pentachlorob ipheny1

Hexachlorobiphenyl

Heptachlorobiphenyl

Octachlorobiphenyl

Nonachlorobiphenyl

Decachlorobiphenyl
  IUPAC
 Congener
 Number
  1

  5

 29

 50

 87

154

188

200

207

209
Chlorine Substitution of
       Congener in
   Calibration Standard
            2

           2,3

          2,4,5

        2,2',4,6

       2,2',3,4,5'

      2,2',4,4',5.6'

    2,2',3,4',5,6,6'

   2,2',3,3',4,5',6,6'

 22'  33'  44'  566'
 £.,£.  , -» , ->  i1*)**  , J i o , o

2,2' ,3,'3',4,4',5,5',$,6'
                                 EXT D-33
                                                                    Rev.  9/88

-------
TABLE D-3,     Quantisation,  Confirmation, and Interference Check
               Ions and PCBs and Internal Standards
Analyte/ Nom. Quant. Confirm.  Expected
   IS
MW
Ion
Ion
Ratio'
             M-70
Accept.  Confirm.
 Ratioa      Ion
 Interference
 Check  Ions
M+70   M+35
FCB Isomer Group
   Clj
   ci
   Cl-

   Clj

   Clt

   Cl
     10
188
222
256
290
324
358
392
426
460
494
188
222
256
292
326
360
394
430
464
498
190
224
258
290
324
362
396
428
466
500
3.0
1.5
1.0
1.3
1.6
1.2
1.0
1.1
1.3
1.1
2.5-3.5
1.3-1.7
0.8-1.2
1.1-1.5
1.4-1.8
1.0-1.4
0.8-1.2
0.9-1.3
1.1-1.5
0.9-1.3
152°
152
186
220
254
288
322
356
390
424
256
292
326
360
394
430
464
498
-
.
222
256
290
326
360
394
430
464
498
.
Internal Standard
         240     240
              241
                5.1
                4.3-5.9
    Ratio of quantitation ion to confirmation ion.

    Monodichlorobiphenyls lost HC1 to produce an ion at m/z 152.
                                 EXT D-34
                                                          Rev.  9/88

-------
TABLE  D-4.
Known  Relative Abundances  of Ions  in PCB Molecular  Ion Clusters
m/z
Relative
Intensity
m/z
Relative
Intensity
Monochlorobiphenyls
188           100
189            13.5
190            33.4
192             4.41

Dichlorobiphenyls
222           100
223            13.5
224            66.0
225             8.82
226            11.2
227             1.44

Tr ichlorob ipheny1s
256           100
257            13.5
258            98.6
259            13.2
260.-           32.7
261             4.31
262             3.73
263             0.47

Te trachlorob iphenyIs
290            76.2
291            10.3
292           100
293            13.4
294            49.4
295             6.57
296            11.0
297             1.43
298             0.95

PentachlorobiphenyIs
324            61.0
325             8.26
326           100
327            13.5
328            65.7
329             8.78
330            21.7
331             2.86
332             3.62
333            0.47
334            0.25
Source:  J.W.  Rote and W.J,
m/z
Relative
Intensity
                   Hexachlorobiphenyls
                   358           50.9
                   359 '           6.89
                   360          100
                   361           13.5
                   362           82.0
                   363           11.0
                   364           36.0
                   365            4.77
                   366            8.92
                   367            1.17
                   368            1.20
                   369            0.15

                   HeptachlorobiphenyIs
                   392           43.7
                   393            5.91
                   394          100
                   395           13.5
                   3.96           98.3
                   397       •    13.2
                   398   .        53.8
                   399,            7.16
                   400           17.7
                   401            2.34
                   402            3.52
                   403            0.46
                   404            0.40

                   Octachlorobiphenyls
                   426           33.4
                   427            4.51
                   428           87.3
                   429           11.8
                   430          100
                   431           13.4
                   432           65.6
                   433            8.76
                   434           26.9
                   435            3.57
                   436            7.10
                   437            0.93
                   438            1.18
                   439            0.15
                   440            0.11
                  Morris,  J. Assoc. Anal.  Chem.  56,  188,  1973.
                             Nonachlorobiphenyls
                             460           26.0
                             461            3.51
                             462           76.4
                             463           10.3
                             464          100
                             465           13.4
                             466           76.4
                             467           10.2
                             468           37.6
                             469            5.00
                             470           12.4
                             471            1.63
                             472            2.72
                             473            0.35
                             474            0.39

                             Decachlorobiphenyls
                             494           20.8
                             495            2.81
                             496           68.0
                             497            9.17
                             498          100
                             499           13.4
                             500           87.3
                             501           11.7
                             502           50.0
                             503            6.67
                             504           19.7
                             505            2.61
                             506            5.40
                             507            0.71
                             508            1.02
                             509            0.13
                                 EXT D-35
                                                         Rev. 9/88

-------
TABLE D-5.
Surrogates
Characteristic Ions for Surrogates and Internal
Standards for Extractable Target Compounds
                Primary  Ion
                                                         Secondary Ion(si
Phenol-dj
2-Fluorophenol
2,4,6-Tribromophenol
Nitrobenzene-dj
2-Fluorob ipheny1
Terphenyl-d^
                    99
                   112
                   330
                    82
                   172
                   244
 42, 71
   64
332, 141
 128,  54
   171
122, 212
Internal Standards
1 , 4- Dichlorobenzene -
Naphthalene - dg
Acenap thene - dg
 Phenanthr ene - d
               i Q
                    152
                    136
                    164
                    188
                    240
                    264
   115
   68
162, 160
 94, 80
120, 236
260, 265
                                  EXT D-36
                                                     Rev! 9/88

-------
                EXHIBIT D
GC/ECD ANALYSES OF AROCLORS AND TOXAPHENE
                                                     Rev.  9/88

-------

-------
1.    Summary of Method

1.1   The analytical method that follows is designed to determine the
      concentration of Aroclors and Toxaphene (Target Compound List [Exhibit
      C]) in high concentration samples.  The method can be used for analyte
      concentrations from the contract required quantitation limits (CRQL) to
      one million times the CRQL in medium and high level waste matrices.

1.2   The Aroclors and Toxaphene listed in Exhibit C are determined by a two-
      column GC/ECD technique.

1.3   This method specifies treatment of extracts with sulfuric acid and
      potassium permanganate.

1.4   GC/ECD analysis begins with initial demonstration of instrument
      performance and calibration of all Aroclors and Toxaphene.  Acceptable
      initial calibration is defined in Section 6.3.  This must be repeated
      whenever the required 12-hour performance evaluation test of Section
      6.4.4 fails or when major instrument maintenance or modification is
      performed.

1.5   Sample extracts must be analyzed within a run sequence as defined in
      Section 6.4.  At a minimum, the sequence consists of an initial
      calibration check using Aroclor standards, method blank analysis,
      sample extract analysis and periodic evaluation mixtures and instrument
      blanks.  (NOTE: Data can only be collected as long as the results for
      the evaluation mixture and instrument blank fall within the limits
      defined in Sections 6.4.3 and 6.4.4.  If two consecutive unacceptable
      evaluation standards are analyzed, all extracts with analytes present
      at >CRQL which have been run since the previous acceptable evaluation
      standard must be reanalyzed.) Additional evaluation mixtures and blanks
      are recommended when highly contaminated samples are suspected.

1.6   Calibration and run sequence specifications for the GC method apply
      separately to both columns.

1.7   One control matrix spike analysis must be run for every 20 single phase
      units in a Case, or once per Case, whichever is more frequent.

1.8   Absolute retention times (RT's) are used for identification of Aroclors
      and Toxaphene.

1.9   The absolute retention time window is calculated from the most recent
      standard as ±1.0 percent of the RT of the standard.

1.10  Aroclors and Toxaphene are identified primarily by pattern recognition,
      but RT's of three to five major peaks muse also be taken into
      consideration.  Guidance on analysis of Aroclors and Toxaphene is given
      in Section 6.7.

1.11  Quantitative analysis of Aroclors and Toxaphene must be accomplished by
      the external standard method as described in Section 6.8.  Three point
      calibration curves for Aroclors and Toxaphene must be generated during
      the initial calibration phase.  A linear response range must be

                                 ARO D-l                              Rev.  9/88

-------
      demonstrated from the CRQL to a high point at least 30 times greater
      than the CRQL.

1.12  Quantitative measurements are made from extracts which have been
      diluted such that ECD response is within the established linear range
      determined by the three-point calibration curve.  Quantitation must be
      performed and reported for both GC columns.

1.13  The surrogates, Tetrachloro-meta-xylene and Decachlorobiphenyl, must be
      added to all samples, blanks, and control matrix spikes analyzed by
      GC/ECD prior to extraction.  The recovery the surrogates will be
      determined in all of these samples and reported to the EPA as a measure
      of method performance.  The retention time shift of the surrogates in
      any standard, sample, control matrix spike,  or blank may not excede
      0.5%.

1.14  Section 6.6 gives criteria which determine whether an analysis is
      complete or whether additional cleanup or dilution is required.

1.15  All samples must be protected from light and refrigerated at 4°C from
      the time of receipt until extraction.

2.    Apparatus and Materials

2.1   Kuderna-Danish (K-D) apparatus.

      2.1.1    Concentrator tube,  10 mL,  graduated (Kontes K-  570040-1029,  or
               equivalent).

      2.1.2    Evaporative  flask,  500 mL (Kontes K-470001-0500, or
               equivalent).

      2.1.3    Snyder column,  three-ball macro  (Kontes K-503000-0121, or
               equivalent).

2.2   Boiling chips.

      2.2.1    Silicon  carbide boiling chips  (optional),  approximately  10-40
               mesh.  Heat  to  400"C for 30  minutes  or  solvent  rinse before
               use.

      2.2.2    Teflon boiling  chips (optional).. Solvent  rinse before use.

2.3   Water bath, heated, with concentric ring cover,  capable of temperature
      control.  NOTE: The bath water should be used in a hood.

2.4   Top loading balance, capable of accurately weighing to +0.01 g.

2.5   Balance-analytical, capable  of accurately weighing to ±0.0001 g.

2.6   Nitrogen evaporation device  equipped with a heated bath that can be
      maintained at  35-40'C, N-Evap by Organomation Associates, Inc., South
      Berlin, MA (or equivalent).
                                  ARO D-2
Rev. 9/88

-------
2.7   Vials and caps, 1 or 2 raL for GC auto sampler.

2.8   Vacuum system for eluting multiple cleanup cartridges.

      2.8.1    Vac Elute Manifold  (Analytichem International,  Harbor City,
               J.T. Baker  or  Supelco)  or  equivalent.

      2.8.2    Vacuum trap made  from  a 500  mL  sidearm flask fitted with a one-
               hole stopper and  glass  tubing.

      2.8.3    Vacuum pressure gauge.

      2.8.4    Rack for holding  10 mL volumetric  flasks  in the manifold.

2.9   Glass vials, at least 20 mL, with screw cap and teflon or inert plastic
      liner for sample extraction and for sulfuric acid and permanganate
      treatment.

2.10  Spatula, stainless  steel or teflon.

2.11  Pipet, Volumetric 1.00 mL or 2.00 mL (optional).

2.12  Syringe, 1.00 mL or 2.00 mL (optional).

2.13  Flask, Volumetric 10.00 mL.

2.14  Flask, Volumetric 1.00 mL or 2.00 mL (optional).

2.15  Vials, 10 mL, with  screw cap and teflon liner  (optional).

2.16  Tube, centrifuge,  12 to 15 mL with 19 mm ground glass joint,
      (optional).

2.17  Snyder Column,  micro two or three ball with a 19 mm ground glass joint.

2.18  Centrifuge, table top (optional)

2.19  Gas chromatographic  system,  including a 0.25 inch injector and an
      electron capture detector.  The GC must be equipped with an integrator
      or data system rather than a strip chart recorder.  Detector makeup gas
      is required for capillary analysis.

      2.19.1  Two wide bore (0.53 mm  ID) fused silica GC columns are
              required.  A separate detector is required for  each column.
              The specified analytical columns are DB-1701, 30 m x 0.53 mm
              ID, 1.0 to  1.5 urn film  thickness, J&W  Scientific, Folsora, CA
              and a DB-608 or SPB-608, 30 m x 0.53 mm ID, 0.8 to 1.5 urn film
              thickness (or equivalent) from J&W Scientific or Supelco, Inc.,
              Bellefonte,  PA.  Equivalent columns may be employed if they
              meet the requirements in 2.19.3.
                                 ARO D-3
Rev. 9/88

-------
      2.19.2  Columns are mounted in 1/4 inch injector ports using glass
              adapters available from a variety of commercial sources (J&W
              Scientific, Supelco, Inc., Hewlett-Packard, Varian, Inc.,
              Perkin Elmer).

      2.19.3  Column equivalence is demonstrated by running the calibration
              standards mixtures described in Section 6.3.1.  Each equivalent
              column must be calibrated according to the procedures described
              in Section 6.3 and satisfy all of the acceptance criteria
              described therein (if an equivalent column is used, it must be
              described in the Case Narrative).

      2.19.4  The carrier gas must be helium.

      2.19.5  Because the column flow for wide bore capillary systems is 5
              mL/minute which is slower than for packed column systems, it is
              necessary to have precise control of the carrier gas flow and
              to supply makeup gas to the detector.  The makeup gas must be
              P-5, P-10  (argon/methane) or nitrogen according to the
              instrument specification.

2.20  Vortex mixer,  Genie,  Model 550-6,  Scientific Industrial,  Inc.,  Bohemia,
      NY (or equivalent).

2.21  Disposable Pasteur pipets,  packed with glass wool rinsed with hexane.

2.22  Ultrasonic cell disruptor,  Heat Systems Ultrasonics,  Inc.,- Model W-385
      SONICATOR (475 Watt with pulsing capability, No.  419 1/8 inch standard
      tapered MICROTIP probe),  or equivalent device with a minimum of 375
      Watt output capability.  NOTE:   In order to ensure that sufficient
      energy is transferred to the sample during extraction, the MICROTIP
      probe must be replaced if the tip begins to erode.  Erosion of the tip
      is evidenced by a rough surface.

2.23  Sonabox acoustic enclosure - recommended with above disrupters for
      decreasing cavitation sound.

3.     Reagents

3.1   Hexane, acetone, iso-octane (optional), and methanol (optional)
      solvents of pesticide residue analysis grade or equivalent.  It is
      recommended that each lot of solvent be analyzed to demonstrate that it
      is free of interference before use.

3.2   Primary Aroclor standards will be obtained from the EPA Quality
      Assurance Materials Bank,  Pesticides and Industrial Chemicals
      Repository, Research Triangle Park,  NC, if available.  Commercial
      standards must be used for all working solutions  after they have been
      shown to be the proper material and at least 95 percent pure by
      comparison to primary standards.

3.3   Mercury (optional).

3.4   Copper powder (optional),  bright and non-oxidized.

                                 ARO D-4                              Rev. 9/88

-------
3.5   Concentrated sulfuric acid (Sp. gr. 1.84).

3.6   Potassium permanganate solution (5 percent w/v).   Slowly add 100 raL
      water to 5 g of potassium permanganate in a Pyrex vessel.

3.7   Diol bonded silica 500-mg, cartridges with stainless steel frits
      Catalog No.  614313, Analytichem, 24201 Frampton Ave., Harbor City, CA
      (or equivalent).

3.8   Ten percent acetone in hexane  (v/v).  Prepare the mixture by adding 10
      mL of acetone to 90.0 mL of hexane.  NOTE:  Prepare this mixture
      accurately or the results from the Diol cartridge cleanup will be
      adversely affected.

3,9   Surrogate Standard Solution.

      The surrogates, tetrachloro-meta-xylene and decachlorobiphenyl, are
      added to all samples, the control matrix spike, and blanks.  Prepare a
      surrogate standard spiking solution of 1.0 ug/mL of each of the
      surrogates in acetone.  The solution must be replaced after six months,
      or sooner, if comparison with quality control check samples indicates a
      problem.

      CAUTION: Analysts must allow all spiking solutions to equilibrate to
      room temperature before use.

3.10  Aroclor control matrix spike solution.  Prepare a spiking solution in
      acetone or methanol that contains 25 ug/mL of Aroclor 1254.              ,

3.11  Performance evaluation standards are a series of 12 mixtures each
      containing the surrogates and one other analyte,  at three concentration
      levels (low, medium, and high).  The other analytes are specified in
      6.4.4.1 (note that the mixture of Aroclors-1016 and 1260 is considered
      as a single analyte).  The concentration level of the low level
      standard is given in 6.3.3.  The medium level standards are 10 times
      the low level concentrations in 6.3.3, and the high level standards are
      30 times the low level concentrations.

4.    Sample Preparation For GC/ECD Analyses

4.1   High concentration samples are initially separated into individual
      phases.   An aliquot of 500 mg of each phase is transferred to a
      separate 20 mL vial.  Wipe the mouth of the vial  with a tissue to
      remove any sample material.  To avoid cross contamination, cap the vial
      before preceding to the next phase unit.

      4.1.1   Add 1.0 mL of the  surrogate spiking solution to  each vial.

      4.1.2   Add 1.0 g of anhydrous sodium sulfate to each vial, and mix
              with a clean spatula.

      4.1.3   Add 9.0 mL of hexane  to each  vial, and sonicate  each vial for  2
              minutes at 100 watts power.   For control matrix  spike, add only
              8.0 mL of hexane  (see  4.3 below).

                                 ARO D-5                              Rev. 9/88

-------
      4.1.4   filter the extract through a disposable pipet loosely packed
              "with glass wool.  Collect at least 7 mL of extract in a clean
              vial.

      4.1.5   Proceed  to paragraph 5.1 for extract cleanup procedures.

4.2   Control matrix spikes, blanks and all single phase units  are spiked
      with the surrogate solution.   The control matrix spike and the blanks
      are subjected to the same extraction, cleanup,  and dilution procedures
      as the samples.

4.3   Once per Case,  or for every group of 20 single phase units (whichever
      is more frequent), a control matrix spike sample is prepared with 1.0
      mL of the Aroclor 1254 spiking solution and 500 mg of corn oil.   The
      control matrix spike must be extracted, cleaned up,  and analyzed in the
      same fashion as  all other single phase units (see 4.1 above).  Note:
      Add only 8.0 mL of hexane to the vial during the extraction of the
      control matrix spike.

4.4   Interferences

      4.4.1   Method interferences may be caused by contaminants in solvents,
              reagents, glassware, and other sample processing hardware;
              these contaminants lead to discrete artifacts, and/or elevated
              baselines, in gas chromatograms.  All of these materials should
              be  routinely demonstrated to be  free from interferences under
              the conditions of the analysis by running reagent blanks.
              Interferences by phthalate esters can pose a major problem  in
              Aroclor  analysis when using  the  electron capture detector.
              Common flexible plastics contain varying amounts of  phthalate
              esters which are easily extracted during laboratory  operations.
              Cross-contamination of clean glassware, routinely occurs when
              plastics are handled.  Interferences from phthalates can best
              be  minimized by avoiding the use of such plastics in the
              laboratory.

      4.4.2   Matrix interferences may be  caused by contaminants that are
              coextracted  from the sample.   The extent of matrix
              interferences will vary considerably from source to  source,
              depending upon the nature of the site being sampled.  The
              cleanup  procedures in Section  5  must be used to remove such
              interferences in order to achieve the contract required
              quantitation limits.

5.    Extract Cleanup  For GC/ECD AnalYses

5.1   Requirements

      5.1.1   This procedure is only suitable  for Aroclors and Toxaphene.
              Because  the  cleanup involves the use of sulfuric acid and
              potassium permanganate, the  final solutions are not  suitable
              for pesticide analysis.
                                  ARO  D-6
Rev. 9/88

-------
      S.I.2   Diol cartridge cleanup is required for all samples, control
              matrix spikes and blanks.  This -procedure removes polar organic
              molecules  such as phenols.  Each lot number of Diol cartridges
              must pass  a cartridge performance check.

      S.I.3   Sulfur can be removed by one of two methods according  to
              laboratory preference (Section 5.6).

      5.1.4   Blanks and control matrix spikes  must be subjected to the same
              cleanup as the field samples.

      5.1.5   It is required that all the waste aliquots be diluted  with
              hexane before initiating the following treatments.

      S.I. 6   No GFC cleanup is used in the preparation of samples for GC/ECD
              analysis under this protocol.

5.2   Sulfuric Acid Cleanup

      5.2.1   Using a syringe or a volumetric pipet, transfer 2.0 mL of the
              hexane solution to a 10 mL vial and carefully add 5 mL of cone.
              sulfuric acid.  This procedure must always be done in  a fume
              hood.  NOTE:  If the remaining extract from the method blank is
              saved at this point, it may be used in paragraph 5.6.3.2 to
              prepare a  sulfur blank.

      5.2.2   CAUTION:  Make sure that there is no exothermic reaction nor
              evolution  of gas prior to proceeding.

      5.2.3   Cap the vial tightly and vortex for one minute.   A vortex must
              be visible in the vial.

      5.2.4   CAUTION:  Stop the vortexing immediately if the vial leaks,
              AVOID CONTACTING THE SOLUTION WITH BARE SKIN,  SULFURIC ACID
              WILL BURN.

      5.2.5   Allow the phases to separate for at least one minute.   Examine
              the top (hexane) layer,  it .should not be highly colored nor
              should it have a visible emulsion or cloudiness.

      5.2.6   If a clean phase separation is achieved proceed to Step 5.2.10.

      5.2.7   If the hexane layer is colored or the emulsion persists for
              several minutes,  remove the sulfuric acid layer from the vial
              via a glass pipette and dispose of it properly.   Add another 5
              mL of clean sulfuric acid.

      5.2.8   Note:  Do  not remove any hexane at this stage of the procedure.

      5.2.9   Vortex the sample and allow the phases to separate as described
              previously.

      5.2.10   Transfer  the hexane layer to a clean 10 mL vial.


                                 ARO D-7                               Rev.  9/88

-------
      5.2.11  Add an additional 1 mL of hexane to the sulfuric acid layer,
              cap the vial securely and shake.  This second extraction is
              done to ensure quantitative transfer of all analytes.

      5.2.12  Remove the second hexane layer and combine with the hexane from
              Step 5.2.10.

5.3   Permanganate Cleanup

      5.3.1   Add 5 mL of the five percent aqueous potassium permanganate
              solution to the combined hexane fractions from 5.2.12.

      5.3.2   CAUTION: Make sure that there is no exothermic reaction nor
              evolution of gas prior to proceeding.

      5.3,3   Cap the vial tightly and vortex for one minute.  A vortex must
              be visible in the vial.

      5.3.4   CAUTION: Stop the vortexing immediately if the vial  leaks,
              AVOID CONTACTING THE SOLUTION WITH BARE SKIN, POTASSIUM
              PERMANGANATE WILL BURN.

      5.3.5   Allow the phases to separate for at least one minute.  Examine
              the top  (hexane) layer, it should not be highly colored nor
              should it have a visible emulsion or cloudiness.

      5.3.6   If a clean phase separation is achieved, proceed to  Step
              5.3.10.

      5.3.7   If the hexane layer is colored or the emulsion persists for
              several minutes, remove the permanganate solution from the vial
              via a glass pipette and dispose of it properly.  Add another  5
              mL of the clean aqueous permanganate solution.

      5.3.8   NOTE: Do not remove any hexane at this stage of the  procedure.

      5.3.9   Vortex the  sample and allow the phases to separate.

      5.3.10  Transfer the hexane layer to a clean 10-mL vial.

      5.3.11  Add an additional 1 mL of hexane to the permanganate layer, cap
              the vial securely and shake.  This second extraction is done  to
              ensure quantitative transfer of all Aroclors and Toxaphene.

      5.3.12  Remove the  second hexane layer and combine with the  hexane from
              Step 5.3.10.

5.4   Final Preparation

      5.4.1   Reduce the volume of the combined hexane layers to 1.0 mL under
              a stream of dry nitrogen.

              5.4.1.1    Nitrogen Slowdown Technique (Taken from ASTM Method
                         D 3086).
                                  ARO D-8
Rev. 9/88

-------
                         5.4.1.1.1    Place the concentrator tube in a
                                      heating bath (30-35'C) and evaporate
                                      the solvent to the final volume using a
                                      gentle stream of clean,  dry nitrogen
                                      (filtered through a column of activated
                                      carbon).   The extract must never be
                                      allowed to become dry.

                         5.4.1.1.2    CAUTION:  New plastic tubing must not be
                                      used between the carbon trap and the
                                      sample, as it may introduce
                                      interferences.   The internal wall of
                                      new tubing must be rinsed several times
                                      with hexane then dried prior to use.

      5.4.2   Prepare the extracts using the Diol cartridge cleanup as
              described below.

5.5   Diol Cartridge Procedure

      5.5.1   Cartridge Performance Check

              Each lot number of Diol cartridges must be tested by the
              following procedure before it  is used for sample cleanup.  Add
              1.0 mL of the control matrix spike solution to 4 mL of hexane.
              Place a 1.0 mL aliquot of the diluted solution onto the top of
              a prewashed Diol cartridge, and elute it with 9 mL of
              hexane/acetone  (90:10(V/V)j.  Adjust the final volume to 10.0
              mL and analyze by .GC/ECD.  The recovery of Aroclor 1254 must be
              determined for evaluation and  reporting purposes..  The lot of
              Diol cartridges is acceptable  if the Aroclor is recovered at 80
              to 110 percent.

      5.5.2   Diol Cartridge Cleanup

              5.5.2.1    Attach the Vac Elute vacuum manifold to a water
                         aspirator or a vacuum pump with a trap installed
                         between the manifold and the vacuum source.   Adjust
                         the vacuum pressure in the manifold to between 5 and
                         10 pounds of vacuum.

              5.5.2.2    A 500 mg Diol cartridge is selected for each hexane
                         solution of waste and placed into the vacuum
                         manifold.

              5.5.2.3    Prior to cleanup of the hexane solutions of waste,
                         the cartridges must be washed with hexane/acetone
                         (90:10).   This is accomplished by placing the
                         cartridge in the  vacuum manifold,  pulling a vacuum
                         and passing 5 mL of the hexane/acetone solution
                         through the cartridge.
                                 ARO D-9                              Rev. 9/88

-------
              5.5.2.4    After the cartridges in the manifold are washed, the
                         vacuum is released and a rack containing labeled 10
                         mL volumetric flasks is placed inside the manifold.
                         Care must be taken to ensure that the solvent line
                         from each cartridge is placed inside of the
                         appropriate volumetric flask as the manifold top is
                         replaced.

              5.5.2.5    After the volumetric flasks are in place, vacuum to
                         the manifold is restored and 1 mL {the entire
                         volume) from each sample, blank, or control matrix
                         spike solution is transferred to the top frit of the
                         appropriate Diol cartridge.

              5.5.2.6    The Aroclors and Toxaphene in the solution
                         concentrates are then eluted through the column with
                         9 mL of hexane/acetone (90:10) and collected into
                         the 10 mL volumetric flasks held in the rack inside
                         of the vacuum manifold.
              5.5.2.7




              5.5.2.8

              5.5.2.9


              5.5.2.10
Transfer the eluate in each volumetric flask to a
clean centrifuge tube or 10 mL vial.  Use two
additional 1 mL hexane rinses of the flask to ensure
quantitative transfer of the cartridge eluate.

Adjust the extract volume to 10.0 mL with hexane

If crystals of sulfur are evident or the presence of
sulfur is suspected, proceed to Section 5.6.

If sulfur is not expected to be a problem, transfer
the 1 mL of solution to a GC vial and label the
vial.  (Some autosamplers require 1 mL solvent
volumes, others require 2 mL.) The solution is ready
for the GC/ECD analysis detailed in Section 6.
Store the extracts at 4°C in the dark until analyses
are performed.
5.6   Sulfur Removal
      5.6.1   Two options are available for the removal of sulfur from
              samples.  The mercury technique appears to be the most
              reliable, but requires the use of small volumes of mercury in
              the laboratory.

      5.6.2   CAUTION: Mercury containing waste should be segregated and
              disposed of properly.

      5.6.3   Mercury Technique

              5.6.3.1    Add 1 to 3 drops of mercury to 1.0 mL of each hexane
                         solution in a' clean vial.  Tighten the top on the
                         vial and agitate it for 30 seconds.   Filter or
                         centrifuge and decant the solution to remove all
                                 ARO D-10
                                             Rev. 9/88

-------
                         solid precipitates and liquid mercury.  Proceed to
                         Section 6 for GC/ECD analysis if the mercury appears
                         shiny.  If the mercury turns black, repeat the
                         process as necessary until it remains shiny.
                         Dispose of the mercury waste properly.

              5.6.3.2    If only a partial set of the hexane concentrates
                         require sulfur cleanup, an additional reagent blank
                         of hexane and mercury (or copper) is required.  This
                         additional blank may be prepared from the remaining
                         blank extract in paragraph 4.14.

      5.6.4   Copper Technique

              5.6.4.1    Bright (non-oxidized) granular copper (one to three
                         granules) can be used in place of mercury in the
                         procedure described in Section 5.6.3.  If the copper
                         appears shiny, proceed to Section 6 for GC/ECD
                         analysis.  If the copper changes color,  repeat
                         sulfur removal as necessary.

6.    GC/ECD Analysis For Aroclors

6.1   Summary

      6.1.1   This GC/ECD method is used for the analysis of the Aroclors and
              Toxaphene only from Exhibit C.  Although this method is similar
              to GC/ECD methods for pesticides, this method is only
              appropriate for Aroclors and Toxaphene.

      6.1.2   1.0 mL of the surrogate spiking solution is added to all single
              phase units, control matrix spikes and blanks prior to
              extraction and surrogate recoveries will be reported with
              sample data.  The retention time shift of the surrogates in any
              standard, sample, control matrix spike, or blank may not excede
              0.5%.

      6.1.3   Control matrix spikes are required for this method and are
              prepared from the Aroclor 1254 control matrix spike solution.

      6.1.4   Quantitation of Aroclors and Toxaphene is done by external
              standard techniques using three-point curves generated during
              an initial calibration sequence.  Quantitation is based on
              comparison of three to five sample peaks with the corresponding
              peaks in the standard.

      6.1.5   The absolute retention times of single component organochlorine
              pesticides are determined as part of this method and reported
              with the sample data in order to prevent misidentification of
              these compounds as constituent peaks of Aroclors or Toxaphene.

      6.1.6   Sample data are collected after an initial calibration sequence
              is run.   The RT's,  calibration factors, and column performance
              are monitored no less than once every 12 hours with an

                                ARO D-ll                              Rev.  9/88

-------
              instrument blank and with a standard.  Data can be collected
              only as long as all evaluation criteria given in Section 6.4.4
            •  are met.  In all GC runs, the injector must be heated to at
              least 205°C.  The injection must be made on column, using
              either automatic or manual injection.  If autoinjectors are
              used, 1.0 uL injection volumes may be used.  Manual injections
              must use at least 2.0 uL injection volumes.

      6.1.7   The carrier gas must be helium.

      6.1.8   The analysis of samples is accomplished by using two wide bore
              (0.53 mm ID) fused silica GC columns.  A separate detector is
              required for each column.  The specified analytical columns are
              DB-1701, 30 m x 0.53 mm ID, 1.0 to 1.5 urn film thickness, J&W
              Scientific, Folsom, CA and a DB-608 or SPB-608, 30 m x 0.53 mm
              ID, 0.8 to 1.5 urn film thickness (or equivalent) from J&W
              Scientific or Supelco, Inc., Beliefonte, PA.  Equivalent
              columns may be employed if they meet the requirements in
              2.19.3.

      6.1.9   Analysis of a sample on both columns is only required when one
              column gives at least one positive result.  Peak identification
              and quantitation must be reported separately for both columns
              if any Aroclors or Toxaphene are detected in a sample.

      6.1.10  Wide bore capillary columns are installed in standard 0.25 inch
              packed column injector and detector ports by use of suitable
              glass adapters and ferrules.  Because the column flow used is 5
              mL/minute of helium, it is necessary to have precise carrier
              gas flow and to supply makeup gas to the detector.

      6.1.11  Electron capture detectors must be plumbed with P-5, P-10
              (argon/methane) or nitrogen as a detector makeup gas according
              to the instrument specification.
      6.1.12  The temperature program for GC analysis is:
                     Initial time
                     Temperature ramp
                     Final hold
                                                         150*C
                                                    1/2 minute
                                            ,5'/minute to 275"C
                                                    10 minutes
      6.1.13
NOTE: It may be necessary to adjust this temperature program
for individual gas chromatographs.   It is a requirement for
this method that the oven(s) regulate temperature.

All calibration and run sequence requirements of the following
sections apply independently to both the specified columns.
6.2   Calibration Standards
      6.2.1   The resolution check standard must be run after the first blank
              in the calibration sequence in order to demonstrate the ability
              of the GC column to resolve certain pesticide compounds.  The
                                 ARO D-12
                                                        Rev. 9/88

-------
               resolution check standard must  contain the  following seven
               compounds each with a concentration of 50 ng/mL:

                          Methoxychlor
                          Endrin ketone
                          p,p'-DDE
                          Dieldrin
                          Endosulfan sulfate
                          Endosulfan I
                          gamma-Chlordane

      6.2.2    The resolution criterion  is  that  the height of the valley
               between two adjacent peaks in the mixture must not be greater
               than 60% of the height of the shorter  peak.   Experience  to date
               suggests that the poorest resolution on the DB-608 column will
               be between DDE and  Dieldrin  and between Methoxychlor and Endrin
               ketone.   On the DB-1701 column, resolution  difficulties  should
               be expected between Endosulfan  I  and gamma-Chlordane and
               between Methoxychlor and  Endosulfan sulfate.

      6.2.3    Each Aroclor must be analyzed at  the three  concentrations given
               in 6.3.1 and 6.3-6.5 in order to  establish  the response  factors
               for the quantitation peaks for  each Aroclor and to demonstrate
               detector linearity.

               The high point concentration defines the upper end of the
               concentration range for which the calibration is valid.

      6.2.4    Aroclor and Toxaphene standards must be  prepared individually,
               except  for Aroclor  1260 and  Aroclor 1016, which can  be combined
               in one  standard mixture.

      6.2.5    The performance of  the GC system  is monitored by the use  of a
               standard and an instrument blank, both of which must be run no
               less than once every 12 hours (see  Section  6.4.4 for the  one
               exception to this requirement).   The instrument blank is  a
               hexane  solution containing 20 ng/mL of each of the surrogates.

6.3   Initial Calibration

      6.3.1    All  columns  in all  GC systems used  to  collect  data using  this
               method must  be calibrated with the  following  sequence of
               standards  (1)  during-the  initial  set up  and  (2) whenever
               corrective action is  required because  a  12-hour check has  shown
               the  instrument to be  out  of  control (Section  6.4.4).
                                 ARO D-13
Rev. 1/89

-------
                                                  No. of Potential
                                         Aroclor Quantitation Peaks
6.3,2
6.3.3
Injection No.   Concentration
              1
              2
              3
              4
              5
              6
              7
              8
              9
             10
             11
             12
             13
             14
             15
             16
             17
             18
             19
             20
             21
             22
             23
             24
             25
             26
Select four or five major peaks from each Aroclor and four
major peaks from Toxaphene as potential quantitation peaks.
The same peaks must be used throughout the run sequence,
including: initial calibration, 12-hour performance checks,
sample quantitation, and control matrix spike analyses.

The low-point concentration calibration standards for the
method analytes are:

               LOW  POINT AROCLOR CALIBRATION  SOLUTIONS
Blank


Resolution Check - —
Low
Medium
High
Low
Medium
High
Low
Medium
High
Low
Medium
High
Low
Medium
High
Low
Medium
High
Low
Medium
High
Low
Low
Blank
1221
1221
1221
1232
1232
1232
1242
1242
1242
1248
1248
1248
1254
1254
1254
1016/1260
1016/1260
1016/1260
Toxaphene
Toxaphene
Toxaphene
Pesticide A
Pesticide B
_
4
4
4
4
4
4
5
5
5
5
5
5
5
5
5
5/5
5/5
5/5
4
4
4
—
—
	
                 Compound

           1.  Aroclor 1016
               Aroclor 1260
           2.  Aroclor 1221
           3.  Aroclor 1232
           4.  Aroclor 1242
           5.  Aroclor 1248
           6.  Aroclor 1254
           7.  Toxaphene
           8.  Tetrachloro-meta-xylene
           9.  Decaclorobiphenyl
                                CAS Number   Low Concentration
                                12674-
                                11096-
                                11104-
                                11141-
                                53469-
                                12672-
                                11097-
                                 8001-
                                  877-
                                 2051-
11-2
82-5
28-2
16-5
21-9
29-6
69-1
35-2
09-8
24-3
 50 ng/mL
 50 ng/mL
 50 ng/mL
 50 ng/mL
 50 ng/mL
 50 ng/mL
 50 ng/mL
500 ng/mL
 20 ng/mL
 20 ng/mL
                           ARO D-14
                                                        Rev. 9/88

-------
6.3.4   The mid-point concentration of each standard is prepared at 10
        times  the concentrations given in 6.3.3.  The surrogates must
        be added at 200 ng/mL to all mid-point calibration solutions.

6.3.5   The high point concentration is prepared at 30-100 times the
        concentrations given in Section 6.3.3.  The surrogates must be
        added  at at least 600 ng/mL to all high point calibration
        solutions.

6.3.6   Determine the absolute retention times (RT) for the selected
        major  peaks for each Aroclor and Toxaphene as well as the
        relative mean deviation (RMD) for each peak,
               HMD -
                     100%  3

                                ^ - RT \ / RT
        where RT is the mean RT for a specific peak in a particular
        Aroclor and Toxaphene.  The RMD for each Aroclor and Toxaphene
        must be less than 0.5 percent before analytical results can be
        reported.

6.3.7   Only three peaks are required for sample quantitation, however,
        the peaks chosen must not co-elute with matrix interference.
        Therefore, linearity of response is required for each of the
        four or five potential quantitation peaks selected during the
        initial calibration in order to establish system performance.

6.3.8   Three-point instrument calibration is required for each
        potential quantitation peak of each Aroclor and Toxaphene'.  The
        laboratory has three choices on how to establish a three point
        calibration.  Only one of the three calibration methods can be
        used to quantitate samples in any single run sequence.
        Laboratories may not mix calibration techniques for samples
        quantitated using a single initial calibration.

        6.3.8.1    The laboratory can use a mean calibration factor
                   (CF) determined from the three concentrations, but
                   only if the % RSD for the three points is < 15.0
                   percent.
                   CF -
         3
         I
       Response of
Selected Peaks in Standard
     Mass Injected (ng)
%RSD -
                          CF
                              x 100
                          ARC D-15
                                             Rev. 9/88

-------
f
                                      Where SD -
                                                   n-1
                                                                          and n-3
                             6.3.8.2
                     6.3.9
                   The laboratory can use a calibration line drawn
                   through all three calibration points if the value
                   for r (correlation coefficient from the linear
                   regression calculation) is >0.975 and if the zero
                   concentration intercept is <0.20.
                           d/n)
                                      Where  r
                                                       n

                                                                       - y)
                              6.3.8.3
                   Laboratories with electronic integrators or data
                   systems that automatically calculate calibration
                   curves as line segments between calibration points
                   may use two line segment calibration curves for each
                   quantitation peak of the Aroclor or Toxaphene.  This
                   technique may be used only if r (the correlation
                   coefficient from the linear regression calculation)
                   is >0.975 for all three points for each peak and if
                   the zero concentration intercept is <0.20 times the
                   low point response for each Aroclor or Toxaphene
                   peak quantitated.

        Mixtures of single component pesticides are injected as part of
        the calibration sequence to establish the RT of individual
        pesticides because they are potential method interferences.
        Calibration factors are not calculated for the individual
        pesticide standards.

6.3.10  Single Component Pesticide Mixtures
Individual
Standard
Mix A
Concentration
(ng/mL)
Individual
Standard
Mix B
Concentrat ion
(ng/mL)
                              alpha-BHC                 25.0
                              Heptachlor                25.0
                              gamma-BHC                 25.0
                              Endosulfan  I              25.0
                              Dieldrin                 50.0
                              Endrin                    50.0
                              p,p'-DDD                 50.0
                              p,p'-DDT                 50.0
                              Methoxychlor              50.0
                              Tetrachloro-meta-xylene   20.0
                              Decachlorobiphenyl        20.0
                                           beta-BHC                25.0
                                           delta-BHC               25.0
                                           Aldrin                  25.0
                                           Heptachlor epoxide      25.0
                                           alpha-Chlordane         25.0
                                           gamma-Chlordane         25.0
                                           p,p'-DDE                50.0
                                           Endosulfan sulfate      50.0
                                           Endrin aldehyde         50.0
                                           Endrin Ketone           50.0
                                           Endosulfan II           50.0
                                           Tetrachloro-meta-xylene 20.0
                                           Decachlorobiphenyl      20.0
                                                ARO D-16
                                                                Rev. 4/89

-------
      6.3.11   The  average  retention times  of  the  surrogates must be
               calculated using all  26  injections  listed  in 6.3.1.  This value
               is used to establish  the acceptance criteria for  all subsequent
               injections.
If (surrogate)
                                26
                           e)  - I  ™  (surrogate)

      6.3.12   The  mean response  for  each  surrogate  in the  low-point
               calibration analysis is used as  the surrogate  calibration
               factor  (CF) .

      6.3.13   Sample  analysis may not proceed  until  a satisfactory
               calibration has been demonstrated.

      6.3.14   The  equivalence of GC  columns other than those specified  in
               2.19.1  may  be  demonstrated  by:

               o  Successful initial  calibration of the gas  chromatographic
                 system including meeting the requirements  in 6.3.6  and
                 6.3.8.

               o  Meeting the resolution criterion in 6.2.2.

               o  Achieving baseline  resolution of each of the components in
                 Individual Mix  A from one another and baseline resolution  of
                 each of the components  in Individual Hix B from one  another.
6.4   Sample Analysis Run Sequence

      6.4.1    Summary

               6.4.1.1   Extracts of single phase units are analyzed as part
                         of a run sequence  that includes both instrument
                         blanks and performance evaluation standards, which
                         provide both calibration verification and instrument
                         performance evaluation.  Both an instrument blank
                         and a standard must be analyzed successfully on each
                         column no less than once in every 12 hours  that
                         sample data are collected.   In addition, the
                         laboratory must analyze at least one method blank
                         for each Case,  or once per 20 single phase  units,
                         whichever is more frequent.  Acceptance criteria for
                         the instrument blanks (Section 6.4.3), method blanks
                         (Section 6.4.2),  and for the performance evaluation
                         standards (Section 6.4.4) are given below.

               6.4.1.2    The laboratory may identify and quantitate  analyte
                         peaks based on data collected during the initial
                         calibration as long as an acceptable instrument
                         blank and an acceptable evaluation mixture  are
                         analyzed every 12 hours.   The requirements  for the
                         run sequence apply to both columns in all
                         instruments  used in these analyses.


                                 ARO  D-17                              Rev.  9/88

-------
        6.4.1.3    Example Run Sequence

                   Note:  The 12 hours are counted from the injection
                   of the sample in step 27, not from step 1.
                   Time
Injection #
Material Injected
                   0 hr.
                   12 hr.
                   24 hr.
1
2
3 - 26

27
o
o
o
o
o
1st injection
  past 12 hr.
2nd injection
o
o
o
o
o
o
1st injection
  past 24 hr.
2nd injection
o
o
Instrument Blank
Resolution Check
Initial Calibration
  Standards and Blanks
First Sample

Subsequent
  Samples

Instrument Blank

Evaluation Mixture
Sample

Subsequent
  Samples
Instrument Blank

Evaluation Mixture
Sample
        6.4.1.4    The run continues until an unacceptable instrument
                   blank or standard is analyzed.  This example run
                   sequence shows only the minimum number of instrument
                   blanks and evaluation mixtures necessary to meet the
                   requirements.   More instrument blanks and
                   evaluations may be run at the discretion of the
                   laboratory, but these must satisfy the criteria
                   presented in Section 6.4.3 and 6.4.4.

        6.4.1.5    A run sequence must include all required control
                   matrix spike analyses and method blanks, but each
                   laboratory may decide at what point in the run
                   sequence they are analyzed.
6.4.2   Method Blanks
        The method blank consists of 500 mg of corn oil spiked with the
        surrogates at 20 ng/mL that is subjected to the same cleanup as
        a sample, in order to check for system contamination.  None pf
        the Aroclors or Toxaphene listed in Exhibit C may be present at
        greater than the CRQL in the analysis of a method blank.  If
        any of the Aroclors or Toxaphene are present at >CRQL, data
                           ARO D-18
                                    Rev.  9/88

-------
         collection must be stopped,  and the data for all  single phase
         units analyzed since the last acceptable blank are  considered
         suspect.   Therefore all single phase units with Aroclors and
         Toxaphene detected at levels >CRQL that were prepared during
         the same  12-hour shift as a  contaminated method blank must be
         reextracted and reanalyzed at no additional  expense to the
         Agency.   At least one method blank must be included for every
         20 single phase units.

6.4.3    Instrument Blank

         An instrument blank is  a hexane solution containing 20 ng/mL of
         the surrogates.   An acceptable instrument blank analysis must
         demonstrate that no Aroclor  or Toxaphene can be detected at
         greater than 0.5 times  the CRQL.   If Aroclors or  Toxaphene are
         detected  at greater than half the CRQL,  all  data  collection
         must be stopped and corrective action taken.   Data  for samples
         with analytes detected  at >CRQL analyzed between  the last
         acceptable instrument blank  and the unacceptable  blank are
         considered suspect.   An acceptable  instrument blank must be
         analyzed  before  additional data are collected.  After an
         acceptable instrument blank  is analyzed,  all  single  phase units
         with Aroclors or Toxaphene detected at  levels >CRQL that were
         analyzed  after the previous  acceptable  instrument blank must be
         reinjected during a valid run sequence  and reported at no
         additional expense to the Agency.

6.4.4    Performance Evaluation  Standards

         6.4.4.1     The performance evaluation standards injected at 12
                   hour  intervals will be rotated so  as to include  the
                    common Aroclors and Toxaphene at low, medium, and
                   high  concentrations according to the following
                    schedule:

                   NOTE:  The time clock starts after  the completion of
                    the initial calibration as in 6.14.3.  The
                   performance evaluation mixture is  injected following
                   an  instrument blank, as in 6.4.13.
                   Concentration

                         low
                         low
                         low
                         low
                       medium
                       medium
                       medium
                       medium
                        high
                        high
                        high
                        high
                       Repeat
Compound

Toxaphene
   1248
   1254
1016/1260
Toxaphene
   1248
   1254
1016/1260
Toxaphene
   1248
   1254
1016/1260
 12
 24
 36
 48
 60
 72
 84
 96
108
120
132
144
                           ARO D-19
                                                                Rev. 9/88

-------
6.4.4.2    For each of the four or five potential quantitation
           peaks (Section 6.3.1),  the RT in the standard must
           be within the retention time window of ±1.0 percent
           of the mean RT calculated during the initial
           calibration.

6.4.4.3    For each of the four or five potential quantitation
           peaks (Section 6.3.1),  the response (area or height)
           must be within 20.0 percent (±20.0 RPD) of the mean
           response obtained during the initial calibration.
RPD -
                                  x 100
                           CF
           Where:
           CF  - Average calibration factor from initial
                 calibration (Equation 6.2).

           CFg - Calibration factor from current
                 evaluation mixture .

6.4.4.4    The retention time shift of the surrogates must be
           within 0.5 percent of the average retention time
           established during the initial calibration.

6.4.4.5    If a performance evaluation standard does not meet
           the criteria listed above, it must be reinjected
           immediately.  If the second injection also does not
           meet the criteria, all data collection must be
           stopped.  Appropriate corrective action must be
           taken and a new initial calibration sequence must be
           analyzed before more sample data are collected.
           Data collected after the last acceptable evaluation
           mixture are considered suspect and all extracts with
           analytes present at > CRQL must be reinjected and
           reported at no additional expense to the Agency.

6.4.4.6    Analysts are cautioned that running an instrument
           blank and an evaluation mixture once every 12 hours
           is the minimum contract requirement.  Highly complex
           samples or unstable GC equipment may cause peaks
           from one injection to be carried over into the next.
           It may be necessary to analyze instrument blanks and
           evaluation mixtures more often in order to avoid
           discarding data.

6.4.4.7    The requirement for running the 12 hour instrument
           blanks and evaluation standards is waived when no
           sample or control matrix spikes are analyzed during
           the 12-hour period.  After a break in sample data
           analysis, a laboratory may resume the analysis of
                   ARO D-20
                                       Rev.  9/88

-------
                          samples and control matrix spikes using the current
                          initial calibration after an acceptable performance
                          evaluation standard is analyzed.  If a successful
                          continuing calibration cannot be demonstrated after
                          an  interruption, an acceptable initial calibration
                          must be run before sample data can be collected.

                          NOTE: This section does not affect the requirement
                          that all acceptable sample analyses must be
                          bracketed by acceptable evaluation mixtures.

      6.4.5    Control Matrix Spike

               6.4.5.1    A control matrix spike must be analyzed once per
                          Case, or once for every 20 single phase units,
                          whichever is more frequent.

                          The percent recovery of each of the five Aroclor
                          1254 peaks is calculated using the following
                          equation:

                          Matrix
                          Spike     -       SSR'SR  x 100
                          Recovery            SA

                          Where:
                          SSR  - Spike Sample Recovery
                        '  SR   - Sample Result
                          SA   - Spike Added

                          These results are reported per Aroclor 1254 peak (5
                          total).

               6.4.5.2     Control matrix spike recoveries will be reported by
                          the Contractor.   (See Exhibit B.)

6.5   Sample Analyses

      6.5.1    The protocol is intended to achieve the quantitation limits
               shown  in Exhibit C whenever possible.  If sample chromatograms
               have interfering peaks, a high baseline, or off-scale peaks,
               then samples must be reanalyzed following dilution, or another
               aliquot  of  the original dilution should be cleaned up and
               analyzed.   Samples which cannot be made to meet specifications
               given  in this section after the second full cleanup (sulfuric
               acid,  permanganate, Diol and mercury sulfur removal) are
               reported as intractable in the Case Narrative and do not
               require  further analysis.  No limit is placed on the number of
               repeat full cleanups of samples that may be required because of
               contaminated method blanks.

      6.5.2    The sample  must be analyzed at the most concentrated level
               consistent  with achieving satisfactory chromatography.  If
               dilution is employed solely to bring a peak within the


                                 ARO  D-21                             Rev. 9/88

-------
              calibration range or to produce an on-scale Aroclor or
              Toxaphene pattern, the results for both a more and a less
              concentrated extract must be reported.  The resulting changes
              in quantitation limits and surrogate recovery must also be
              reported for the dilute samples.

      6.5.3    If the laboratory has reason to believe that diluting the final
              extracts will be necessary, an undiluted run may not be
              required.  If an acceptable chroraatogram (as defined in 6.6) is
              achieved with the diluted cleaned sample extract, an additional
              analysis at 10 times the concentration of the dilution must be
              injected and reported with the sample data.

      6.5.4    The peak response of sample peaks in diluted cleaned samples
              must be >25 percent of full scale to allow visual pattern
              recognition of the Aroclors and Toxaphene.

      6.5.5    An on-scale chromatogram(s) of all selected Aroclor
              quantitation peaks must be presented with the sample data.

      6.5.6    Chromatographic data may be replotted electronically in order
              to produce an on-scale chromatogram, except when the off-scale
              sample peaks are larger than the high point calibration peaks.
              In that case, the samples must be diluted and reinjected.

      6.5.7    The peak response of sample peaks on the replotted chromatogram
              must be >25 percent of full scale to allow visual pattern
              recognition of the Aroclors or Toxaphene.

6.6   Data Acceptance Criteria

      6.6.1    Reportable data for a sample must include a chromatogram with a
              baseline which returns to below 50 percent of full scale before
              the elution time of Aroclor 1221 and to below 25 percent of
              full scale after Aroclor 1221 and before the elution time of
              Decachlorobiphenyl.

      6.6.2    If dilution has been applied and no peaks are detected above 25
              percent of full scale, analysis of a more concentrated sample
              is required.

      6.6.3    Reportable sample data must include chromatogram(s) with all
              detected Aroclor or Toxaphene peaks in the linear range of the
              quantitation^determined by the initial calibration.

      6.6.4    NOTE: If more than one chromatogram is required to satisfy the
              criteria for a sample, the results of all chromatograms must be
              reported.  These requirements apply to both columns.

      6.6.5    Peaks used for quantitation that are more than two times the
              width of the high concentration calibration peaks for that
              analyte must be reported with an appropriate data flag.  (See
              Exhibit B.)
                                 ARO D-22
Rev. 9/88

-------
      6.6.6   The  retention time shift of the surrogates in any standard,
              sample, control matrix spike, performance evaluation standard,
              or blank may not excede 0.5%.

6.7   Identification Of Aroclors And Toxaphene

      6.7.1   Aroclors and Toxaphene present special analytical difficulties.
              Because of  the alteration of these materials in the
              environment, it is probable that samples which contain Aroclors
              or Toxaphene will give similar but not identical patterns as
              Aroclor or  Toxaphene standards.  Thus, identification requires
              visual inspection of an on-scale pattern.  The pattern may be
              brought on-scale either by diluting the sample and reinjecting,
              or by replotting the chromatograph stored in a laboratory data
              system.

      6.7.2   The  choice  of the peaks used for quantitation and recognition
              of those peaks may be complicated by the environmental
              alteration  of the Aroclors and Toxaphene and by the presence of
              coeluting analytes and/or matrix interferences.

      6.7.3   The  numbers of potential quantitation peaks are listed in
              6.3.1.

      6.7.4   The  more highly chlorinated components of the Aroclors and
              Toxaphene are more stable in the environment.  Therefore, the
              analyst should emphasize the later eluting peaks of a pattern
              in identifying analytes when weathered Aroclors or Toxaphene
              are  detected.     •                       _

6.8   Quantitation Of Analytes

      6.8.1   Analytes may be quantitated using either a modern electronic
              integrator  or a laboratory data system.  The analyst may use
              either peak height or peak area as the basis for quantitation,
              however the use of area versus height must be consistent
              between samples.   The use of an electronic integrator or a
              laboratory data system is required.

      6.8.2   The chromatograms of all samples must be reviewed by a
              qualified Pesticide/PCB analyst before they are reported.

      6.8.3   Using an electronic integrator,  one of three calibration
              techniques may be employed (6.3.5).   It is the responsibility
              of the analyst to set the integration parameters such that off-
              scale chromatograms are within the dynamic range of the
              instrument.   The  analyst should also check for data flags
              generated by the  instrument that indicate improper quantitation
              of peaks prior to reporting data to the EPA.

      6.8.4   In order to be quantitated,  the  detector response (peak area or
              peak height) of all analytes  must lie between.the responses  of
              the low and high  concentrations  in the initial calibration.   If
              the analytes are  detected below  the  CRQL,  they are reported

                                ARO D-23                              Rev. 9/88

-------
6.8.6
6.8.7
6.8.8
6.8.9
6.8.10
6.8.11
6.8.12
with the appropriate flags (See Exhibit B).   If they are
detected at a level greater than the high calibration point,
the sample must be diluted either to a maximum of 1:100,000, or
until the response is within the linear range established
during calibration.

The analyst must select 3-5 quantitation peaks for each
detected Aroclor or Toxaphene that do not co-chromatograph with
matrix interferences and determine the concentration using each
peak separately.  Concentrations are calculated using the
following equations:

Samples:

       Concentration mg/Kg —  	
                                           (AX)
        Where:
        OF -
        W,
           Response for the parameter to be measured (height or
           area).

           Calibration factor for the external standard
           (Section 6.3.5).

           Volume of total extract (uL) (take into account any
           extra dilution).

           Volume of extract injected (uL).

           Weight of waste diluted (mL).
The laboratory will quantitate each of the selected Aroclor or
Toxaphene peaks individually, and determine an average
concentration from all of the selected peaks.  Quantitation is
performed on both columns; the lower value is reported on Form
I.  See Exhibit B for instructions on completing Form I and
Form X.

Detected Aroclors and Toxaphene must be reported as: (1) a
concentration between the CRQL and 10  x CRQL, or (2) as an
estimated value below the CRQL.  (See Exhibit B.)

If more than one Aroclor is observed in a sample or if an
Aroclor and Toxaphene are observed in the same sample, the
laboratory must choose separate congener peaks to quantitate
the different analytes.  A peak common to both analytes present
in the sample must not be used to quantitate both analytes.

The concentrations of the surrogates are calculated using the
equation in 6.8.7, where the mean response of the surrogate
established during the initial calibration serves as the
surrogate calibration factor (CF).

The retention time shift of the surrogates in any standard,
sample, control matrix spike, or blank may not excede 0.5%.
                           ARO D-24
                                                        Rev. 4/89

-------
                   EXHIBIT E
Quality Assurance/Quality Control Requirements
                                                    Rev.  9/88

-------

-------
                              TABLE OF CONTENTS
Section
Page Number
I.     Introduction	E-1




II.     QA/QC Standard Operating Procedures	E-2




III.   QA/QC Requirements




       Volatile QA/QC Requirements	VOA E-8




       Extractables QA/QC Requirements	EXT E- 25




       Aroclors/Toxaphene QA/QC Requirements	ARO E-45




IV.     Analytical Standards	E-63




V.     Laboratory Evaluation Procedures	E-65
                                                                    Rev. 9/88

-------

-------
                QUALITY ASSURANCE/QUALITY CONTROL REQUIREMENTS
                                  SECTION I
                            INTRODUCTION AND SCOPE
      The Quality Assurance/Quality Control (QA/QC) procedures defined herein
must be used by the Contractor when performing analyses according to the
methods specified in Exhibit D.  This exhibit summarizes the QA/QC procedures
and criteria that are mandatory for the performance of the Contract.

      The purpose of this document is to provide a uniform set of procedures
for the analysis of high-concentration organic samples, documentation of
methods and their performance, and verification of the sample data generated.
The program will also assist laboratory personnel in recalling and defending
their actions under cross examination if required to present court testimony
in enforcement case litigation.
                                   E-l                         •    Rev. 9/88


-------
             SECTION  II
QA/QC STANDARD OPERATING PROCEDURES
              E-2
Rev. 9/88

-------
1.    General OA/OC Consideration

      The Contractor shall have a written QA/QC SOP (Standard Operating
      Procedure) which describes the inhouse procedures that he employs to
      guarantee, to the extent possible, the quality of all analysis
      activities.  It should describe the quality assurance and the quality
      control procedures used during the analysis.   Each Contractor should
      prepare his own SOPs to suit the needs of his organization as he has
      best determined.  The QA/QC SOP should contain the essential elements
      described in this section.

2.    Elements of a QA/OC SOP

2.1   All routine laboratory tasks should have written QA/QC Standard
      Operating Procedures.  Standard Operating Procedures should be detailed
      documents describing who does what, when, where, how, and why.  They
      shall be sufficiently complete and detailed to ensure:

      2.1.1    Data of known  quality  and integrity are  generated.

      2.1.2    The loss of  data  due to out of control  conditions is  minimized.

2.2   Standard Operating Procedures shall be:

      2.2.1    Adequate to  establish  the traceability  of standards,
               instrumentation,  samples,  and environmental  data.

      2.2.2    Simple, so a user'with basic  education,  experience  and/or
               training can properly  use them.

      2.2.3    Complete enough so  the user follows the  directions  in a
               stepwise manner.

      2.2.4    Consistent with sound  scientific principles.

      2.2.5    Consistent with current-EPA regulations, guidelines,  and
               contract requirements.

      2.2.6    Consistent with the instrument manufacturer's specific
               instruction  manuals.

2.3   Standard Operating Procedures will also  provide for documentation
      sufficiently complete to:

      2.3.1   Record  the performance of all tasks and  their results.

      2.3.2   Explain the  cause of missing data.

      2.3.3   Demonstrate  the validation of data each  time they are recorded,
              calculated,  or transcribed.
                                   E-3
Rev. 9/88

-------
2.4   To accomplish these objectives,  Standard Operating Procedures should
      address the major elements upon which the final quality of the
      contractors work depends.   In the following descriptions these six
      major areas have been divided into subelements, where applicable.
      These elements include but are not limited to:

      2.4.1   Organization and personnel,

      2.4.2.  Facilities and equipment,

      2.4.3.  Analytical methodology,

      2.4.4   Sample custody procedures,

      2.4.5   Quality control, and

      2.4.6   Data handling.

3.   Organization and Personnel

3.1   QA Policy and Objectives - Each organization should have a written
      quality assurance policy that should be made known to all organization
      personnel.  Objectives should be established to produce data that meet
      contract requirements in terms of completeness, precision, accuracy,
      representativeness, documentation, and comparability.  1'he SOP should
      require the preparation of a specific QA plan for the analysis.

3.2   QA Organization - The organization and management of the QA function
      should be described in the Contractor's SOP.  Reporting relationships
      and responsibilities should be clearly defined.  A QA Coordinator or
      Supervisor should be appointed and his responsibilities established.  A
      description of the QC paperwork flow should be available.  There should
      be a clear designation of those who are authorized to approve data and
      results.  Responsibilities for taking corrective action should be
      assigned to appropriate management personnel.

3.3   Personnel Training - It is highly desirable that there be a training
      program for employees.  This system should include motivation toward
      producing data of acceptable quality and should involve "practice work"
      by the new employee.  The quality of this work can be immediately
      verified and discussed by the supervisor, with appropriate corrective
      action taken.

3.4   Document Control and Revisions - The SOP should include a system for
      documenting:

      3.4.1   Calibration procedures,

      3.4.2   Analytical procedures,

      3.4.3   Computational procedures,

      3.4.4   Quality control procedures,
                                   E-4
Rev. 9/88

-------
      3.4.5    Bench data,

      3.4.6    Operating procedures,  or  any  changes  to  these procedures,  and

      3'.4.7    Laboratory notebook policy.

3.5   Procedures making revisions to technical procedure or documents must be
      clearly defined, with the lines of authority indicated.   Procedural
      revisions should be written and distributed to all affected  -
      individuals,  thus ensuring implementation of changes.

4.    Facilities and Equipment

4.1   Procurement and Inventory Procedures - Purchasing guidelines for all
      equipment and reagents having an effect on data quality should be well
      defined and documented.  Similarly, performance specifications should
      be documented for all items of equipment having an effect on data
      quality.  Once any item which is critical to the analysis such as an in
      situ instrument, or reagent is received and accepted by the
      organization, documentation should be retained of the type, age, and
      acceptance status of the item.  Reagents should be dated upon receipt
      in order to establish their order of use and to minimize the
      possibility of exceeding their useful shelf life.

4.2   Preventive Maintenance - Preventive maintenance procedures should be
      clearly defined and written for each measurement system and required
      support equipment.  When maintenance activity is necessary, it should
      be documented on standard forms maintained in logbooks.   A history of
      the maintenance record of each system serves as an indication of the
      adequacy of maintenance schedules and parts inventory.

5-    Analytical Methodology

5.1   Calibration and Operating Procedures - Calibration is the process of
      establishing the relationship of a measurement system output to a known
      stimulus.  In essence, calibration is a reproducible reference point to
      which all sample measurements can be correlated.   A sound calibration
      SOP should include provisions for documentation of frequency,
      conditions,  standards, and records reflecting the calibration history
      of a measurement system.

      5.1.1   The accuracy of the calibration standards is an important point
               to consider since all data will be  in  reference to the
              standards used,  An SOP for verifying  the accuracy of all
              working standards against primary grade standards  should be
              routinely followed.

5.2   Feedback and Corrective Action - The SOP should specify the corrective
      action that is to be  taken when an analytical or sampling error is
      discovered or the analytical system is determined to be  out of control.
      The SOP should require documentation of the corrective action and
      notification of the analyst of the error and correct procedures.
                                   E-5
Rev. 9/88

-------
6.   Sample Custody

6.1   Sample custody Is a part of any good laboratory or field operation.
      Where samples may be needed for legal purposes, "chain of custody"
      procedures, as defined in Exhibit F must be used.  However, at a
      minimum, the following sample custody procedures should be addressed in
      the QA/QC SOP.

6.2   Chain of Custody in Laboratory Operations

      6.2.1    Identification of  responsible party  to act as  sample  custodian
               at the  laboratory  facility  authorized  to sign  for  incoming
               field samples, obtain  documents  of shipment  (e.g., bill  of
               lading  number or mail  receipt),  and  verify the data entered
               onto the  sample custody  records.

      6.2.2    Provision for a laboratory  sample  custody log  consisting of
               serially  numbered  standard  lab  tracking report sheets.

      6.2.3    Specification of laboratory sample custody procedures for
               sample  handling, storage and dispersement of analysis.

7.    Quality Control

7.1   Quality Control Procedures - The quality control procedures used during
      analysis should be described-and must conform  to those described in
      Exhibit E.  The quality control checks routinely performed during
      sample analysis include reagent blank analysis to establish analyte
      levels, control matrix spike and blank sample  analysis to determine
      analytical accuracy.  The frequency of these qualitjFlissurance checks
      are defined  in the contract.  Limits of acceptance or rejection are
      also defined for analysis and control charts should be used.
      Confirmation procedures should be described in the SOP.

7.2   Control Checks and Internal Audits •? A good SOP will make provision for
      and describe control checks and internal audits by the Contractor.
      Several approaches are used for control checks.  These include:

      7.2.1    Reference Material Analysis  - Analytical reference-'materials
               are available from several  commercial  and government  sources,
               or they may be prepared  inhouse.   The  chemical analysis  of
               these materials has been well established.   Such materials can
               be analyzed alongside  routine samples  and the  results used to
               check the accuracy of  analytical procedures.

      7.2.2    Blank Analysis - The procedures  and  the  frequency  of  blank
               analyses.are defined in  the  contract.

      7.2.3    Control Matrix Spike Analysis -  The  procedures and the
               frequency of matrix spike analyses are defined in  the contract.
                                    E-6
Rev. 9/88

-------
      7.2.4   Internal Audits - Internal audits should be periodically
              conducted to evaluate the functioning of the QA SOP.  This
              involves an independent check of the performance of the
              laboratory analysts to determine if prescribed procedures are
              closely followed.

8.    Data Handling

8.1   Data Handling, Reporting,  and Recordkeeping -  Data handling,  reporting,
      and recordkeeping procedures should be described.   Data handling and
      reporting includes all procedures used to record data on standard
      forms, and in laboratory notebooks.   The reporting format for different
      types of bench data should be described and the forms provided.   The
      contents of notebooks should be specified.

      8.1.1   Recordkeeping of this type serves at least two useful
              functions: (1) it makes possible the reanalysis of a set of
              data  at a future time, and (2) it may be used in support of the
              experimental conclusions if various aspects of the analysis are
              called into question.

8.2   Data Validation -  Data validation procedures,  defined ideally as a set
      of computerized and manual checks applied at various  appropriate levels
      of the measurement process,  should be in written form and clearly
      defined for all measurement systems.

      8.2.1   Criteria.for data validation must-be documented and include
              limits on:

              8.2.1.1    Operational parameters such as GC conditions;

              8.2.1.2    Calibration data;

              8.2.1.3    Special checks unique to each measurement,  e.g.,
                         successive values/averages;

              8.2.1.4    Statistical tests,  e.g.  outliers;  and

              8.2.1.5    Manual checks such as hand  calculations.

      8.2.2   The limits defined in the contract ensure a high probability of
              detecting invalid data for either all  or the majority of
              activities.  (GC operating conditions,  analytical precision,
              etc. should be recorded on standard forms in a logbook.)
                                   E-7
Rev. 9/88

-------
         SECTION  III
VOLATILES QA/QC REQUIREMENTS
         VOA  E-8                               Rev.  9/88

-------
                             TABLE OF CONTENTS
PART 1 - Tuning and GC/MS Mass Calibration






Part 2 - Calibration of the GC/MS System






Part 3 - Method Blank Analysis






Part 4 - Surrogate Spike (SS) Analysis






Part 5 - Control Matrix Spike Analysis






Part 6 - Sample Analysis
Page Number




 VGA E-10







 VOA E-ll






 VGA E-16







 VOA E-17






 VOA E-19






 VOA E-20
                                 VOA E-9
                Rev. 9/88

-------
This section outlines the minimum quality control (QC) operations necessary
to satisfy the analytical requirements associated with the determination of
volatile organic target compounds in waste samples.  These QC operations are
as follows:
      o Documentation of GC/MS Mass Calibration  and Abundance Pattern

      o, Documentation of GC/MS Response  Factor Stability

      o Internal Standard Response and Retention Time Monitoring

      o Method Blank Analysis

      o Surrogate  Spike Response Monitoring

      o Control Matrix Spike Analysis

PART 1  -  TUNING AND GC/MS MASS CALIBRATION

1.

      It is necessary to establish that a given GC/MS meets the standard mass
      spectral abundance criteria prior to initiating any on-going data
      collection.  This is accomplished through the analysis of
      p-Bromofluorobenzene (BFB).

      Definition:  The twelve (12) hour time period for GC/MS system tuning
      and standards  calibration (initial or continuing calibration criteria)
      begins at the moment of injection of the BFB analysis that the
      laboratory submits as documentation of a compliant tune.  The time
      period ends after twelve (12.0) hours has elapsed according to the
      system clock.

1.1   p-Bromofluorobenzene (BFB)

      1.1.1    Each GC/MS system  used  for the  analysis of volatile  target
               compounds must be  hardware tuned to  meet  the  abundance  criteria
               listed in Table  1.1  for a  maximum  of a  50 nanogram injection of
               BFB.   Alternately, add  50  ng of BFB  solution  to 5.0 mL  of
               reagent water and  analyze  according  to  Exhibit D VOA, Section
               IV.   This criterion  must be demonstrated  daily or  for each
               twelve (12) hour time period, whichever is more frequent.  If
               required, background subtraction should be straightforward and
               designed only to eliminate column  bleed or instrument
               background ions.   Background subtraction actions that result in
               spectral distortions for the purpose of meeting the contract
               specifications are unacceptable.

               NOTE:  All instrument conditions must be identical to those
               used in sample analysis, except that a different temperature
               program will be used.

      1.1.2    BFB  criteria MUST be met before any standards, samples or
               blanks are analyzed.  Any  samples  analyzed when tuning criteria
               have  not been met may require reanalysis at no additional cost
               to the Agency.
                                 VOA E-10
Rev. 9/88

-------
      1,1.3    Whenever the  Contractor  takes  corrective action which may
               change  or affect  the  tuning  criteria  for BFB  (e.g.,  ion
               source  cleaning or  repair, etc.),  the tune must be verified
               irrespective  of the 12-hour  tuning requirements.

                          TABLE 1.1 BFB KEY IONS AND ABUNDANCE CRITERIA

                      Mass            Ion Abundance Criteria

                        50       15.0-40.0 percent of mass 95
                        75       30.0-60.0 percent of mass 95
                        95       Base peak, 100 percent relative abundance
                        96       5.0-9.0 percent of mass 95
                       173       Less than 2.00 percent of mass 95
                       174       Greater than 50.0 percent of mass 95
                       175       5.0  - 9.0 percent of mass 174
                       176       Greater than 95.0 percent but less than
                                101.0 percent of mass 174
                       177       5.0  - 9.0 percent of mass 176
1.2   Documentation

      The Contractor shall provide documentation of the calibration in the
      form of a bar graph spectrum and as a mass listing.

      1.2.1   The Contractor shall complete a Form V  (GC/MS Tuning and Mass
              Calibration) each time an analytical system is tuned.  In
              addition,  all standards, samples, control matrix spikes and
              blanks analyzed during a particular tune must be summarized
              in chronological order on the bottom of the appropriate Form
              V.  Detailed instructions for the completion of Form V are in
              Exhibit B, Section  III.

PART 2 - CALIBRATION OF THE GC/MS SYSTEM

2.    Summary

      Prior to the analysis of samples,  method blanks or control  matrix
      spikes, and after tuning criteria have been met,  the GC/MS  system
      must be initially calibrated at a minimum of five concentrations  to
      determine  the linearity of response utilizing target compound
      standards.   Once  the system has been calibrated,  the calibration must
      be verified each  twelve (12) hour time period for each  GC/MS system.
                                 VOA E-ll                              Rev. 9/88

-------
2.1.  Prepare calibration standards as described in Exhibit D VOA, Section
      IV, to yield the following specific concentrations:

      2.1.1   Initial calibration of volatile target compounds  is required
              at 20, 50, 100,' 150 and 200 ug/L in water.   (This corresponds
              to 10, 25, 50,  75, and 100 mg/kg in a 1 g sample,)  Add
              working standards of the target compounds (in methanol) to
              reagent water.  Add additional methanol so that the total
              volume of methanol is 100 uL per 5 mL of aqueous  standard.
              Utilizing the analytical protocol specified  in Exhibit D,
              this will result in 100-1000 total ng analyzed.   If a
              standard analyte saturates at the 200 ug/L concentration
              level, and the  GC/MS system is calibrated to achieve a
              detection sensitivity of equal to or greater than 5 ug/L (in
              water), the Contractor must document it in the Case
              Narrative, and  proceed with a four-point initial  calibration
              for that specific analyte.

2.2   The USEPA plans to develop performance-based criteria for response
      factor data acquired during this program.   To accomplish this goal,
      the Agency has specified the concentration levels for initial
      calibration and has also specified the specific internal standard to
      be used on a compound-by-compound basis for quantitation (see Table
      2.1).  Establishment of standard calibration procedures is necessary
      and deviations by Contractors will not be allowed.

2.3   Analyze each calibration standard and tabulate the area of the
      primary characteristic  ion (Exhibit D VOA,  Table 3)  against
      concentration for each compound including all contract required
      surrogate compounds.   The relative retention times of each compound
      in each calibration run should agree within 0.06 relative retention
      time units.  Late eluting compounds usually will have much better
      agreement.

            Using Table 2.1 and Equation 2.1, calculate the relative    ^
            response factors  (RRF) for each compound at each concentration
            level.
                 RRF  -
cis
Equation 2.1
                           Us
            where,
            AX  -   Area of the characteristic ion for the
                    compound to be measured.
            A^s -   Area of the characteristic ion for the
                    specific internal standards from Table 3 in
                    Exhibit D.
            C.  -   Concentration of the internal standard
                    (ng/uL).

            C   -   Concentration of the compound to be measured
                    (ng/uL).
                                 VOA E-12
                                   Rev.  9/88

-------
            TABLE 2.1  VOLATILE INTERNAL STANDARDS WITH CORRESPONDING
                    TARGET ANALYZES ASSIGNED FOR QUANTITATION
Bromochloromethane
1,4-Difluorobenzene
             Chlorobenzene-de
Chioromethane
Bromomethane
Vinyl Chloride
Chloroethane
Methylene Chloride
Acetone
Carbon Disulfide
1,1-Dichloroethene
1,1-Dichloroethane
1,2-Dichloroethene
  (total)
Chloroform
1,2-Dichloroethane
2-flutanone
1,2-Dichloroethane-d^
  (surr)
1,1,1-Trichloroethane
Carbon Tetrachloride
Vinyl Acetate
Bromodichloromethane
1,2-Dichloropropane
trans-1,3 -Dichloropropene
Trichloroethene
Dibromochloromethane
1,1,2-Trichloroethane
Benzene

cis-l,3-Dichloropropene
Bromoform
             2-Hexanone
             4-Methyl- 2 -Pentanone
             Tetrachloroethene
             1,1,2,2-Tetrachloroethane
             Toluene
             Chlorobenzene
             Ethylbenzene
             Styrene
             Total Xylenes
             Bromofluorobenzene
               (surr)

             Toluene-.dg (surr)
(surr) - surrogate compound
      2.3.1   Using the average relative response factors  (RRF) from the
              initial calibration, calculate the percent relative standard
              deviations  (%RSD) for compounds labeled on Form VI as
              Calibration Check Compounds  (CCC) and shown  in Table 2.2  (see
        -  -  '  2.6.2) using Equation 2.2 below.  The calibration check
              compounds for volatiles are:  vinyl chloride, 1,1-
              dichloroethene, chloroform,  1,2-dichloropropane, toluene, and
              ethylbenzene.
                     %RSD  -
       SD
        X
100
Equation 2.2
                     where,
                     RSD -  Relative Standard Deviation
                     SD  -  Standard Deviation of 5 initial
                            response factors (per compound)
                                 VGA E-13
                                             Rev. 9/88

-------
                     where,    SD  -  A /.Y    (xt -  X)2
                                                 n-1

                     x  -  mean of 5 initial response factors (per
                           compound)

                     The %RSD for each individual Calibration Check
                     Compound must be less than or equal to 30.0 percent.
                     This criterion must be met for the initial calibration
                     to be valid.

2.4   A system performance check must be performed to ensure that minimum
      average relative response factors are met before the calibration
      curve is used.

            2.4.1    For volatiles, the five System Performance Check
                     Compounds (SPCC) are:  chloromethane, 1,1-
                     dichloroethane, bromoform, 1,1,2,2-tetrachloroethane,
                     and chlorobenzene.  The minimum acceptable average
                     relative response factor (RRF)  for these compounds is
                     0.300 (0.250 for Bromoform).  These compounds
                     typically have RRFs of 0.4-0.6 and are used to check
                     compound instability and check for degradation caused
                     by contaminated lines or active sites in the system.
                     For instance:

                     o   Chloromethane  -  this  compound  is  the  most  likely
                         compound to be lost if the  purge  flow is  too  fast.

                     o   Bromoform -  this compound is one  of the compounds
                         most likely to be purged very  poorly  if the purge
                         flow is  too slow.   Cold spots  and/or  active sites
                         in the  transfer  lines may adversely affect
                         response.   Response of the  quantitation ion (m/z
                         173)  is  directly affected by the  tuning of BFB  at
                         ions m/z 174/176.   Increasing  the m/z 174/176 ratio
                         may  improve bromoform response.

                     o   Tetrachloroethane and 1,1-Dichloroethane  - These
                         compounds can  be deteriorated  by  contaminated
                         transfer lines in purge and trap  systems and/or
                         active  sites in  trapping materials.

      2.4.2   The  initial calibration  is  valid  only after the criteria  for
              the  %RSD  for the CCC  and the  minimum average  RRF for the  SPCC
              have both been met.   Only  after both of  these criteria are
              met  can sample  analysis  begin.

2.5   Documentation

      Once the initial calibration is validated, calculate and report the
      average relative response factor (RRF) and percent relative standard
      deviation (%RSD) for all target compounds.  The Contractor shall

                                 VOA E-14                             Rev..9/88

-------
      complete and  submit Form V (the CC/MS tune for the initial
      calibration) and Form VI (Initial Calibration Data) for each
      instrument used to analyze samples under this protocol.  Detailed
      instructions for completion of Forms V and VI are found in Exhibit B,
      Section III.

2.6   Continuing Calibration

      A calibration standard(s) containing all volatile target compounds,
      including all required surrogates, must be performed each twelve (12)
      hours during analysis (see definition of twelve-hour time period,
      paragraph 1 of this Section).   Compare the relative response factor
      data from the standards run each twelve hours with the average
      relative response factor from the initial calibration for a specific
      instrument.  A system performance check must be made each twelve
      hours.  If the SFCC criteria are met, a comparison of relative
      response factors is made for all compounds.  This is the same check
      that is applied during the initial calibration (Form VI).  If the
      minimum relative response factors are not met, the system must be
      evaluated and corrective action must be taken before sample analysis
      begins.

      2.6.1   Some possible problems are standard mixture degradation,
              injection port  inlet contamination, contamination at the
              front end of the analytical  column, and active sites in the
              column or chromatography  system.  This check must be met
              before analysis begins.   The minimum relative response factor
              (RRF) for volatile System Performance Check Compounds (SPCC)
              is 0.300 (0.250 for Bromoform).

      2.6.2   Calibration Check Compounds  (CCC)

                     After the system performance check is met,  Calibration
                     Check Compounds  (CCC)-listed in Table 2.2  are used to
                     check the validity of the initial  calibration.
                     Calculate the percent difference using Equation 2.3.
                  .   %  Difference   -  RRFT  -  RRFC               Equation 2.3
                                        RRFj
                    where,
                    RRFj -  average  relative  response  factor  from initial
                            calibration.
                    RRFC -  relative response factor from  current
                            verification  check standard.

              2.6.2.1    If  the percent difference  for any compound is
                         greater than 20%,  the Contractor  should consider
                         this  a warning limit.  If  the percent difference
                         for each CCC is  less than  or  equal to 25.0%,  the
                         initial calibration  is assumed to be valid.   If

                                 VOA E-15                              Rev. 9/88

-------
                         the criteria are not met (>25.0% difference), for
                         any one calibration check compound, corrective.
                         action MUST be taken.  Problems similar to those
                         listed under SPCC could affect these criteria.  If
                         no source of the problem can be determined after
                         corrective action has been taken, a new initial
                         five point calibration MUST be generated.  These
                         criteria MUST be met before sample analysis
                         begins.

                                TABLE 2.2  VOLATILE CALIBRATION CHECK
                                           COMPOUNDS (CCC)

                                      1,1-Dichloroethene
                                      Chloroform
                                      1,2-Dichloropropane
                                      Toluene
                                      Ethylbenzene
                                      Vinyl Chloride
      2.6.3   Concentration Levels  for Continuing Calibration Check

              The USEPA plans  to evaluate  the long  term stability of
              relative response factors during this program.
              Standardization  among contract laboratories  is necessary  to
              reach  these  long term goals.  Along with contract  specified
              concentrations for initial calibration, the  USEPA  is
              requiring specific concentrations for each continuing
              calibration  standard(s).
               2.6.3.1
2.7   Documentation
The concentration for each volatile target
compound in the continuing calibration standard(s)
is 50 ug/L.
            The contractor shall complete and submit a Form VII for each
            GC/MS system utilized for each twelve hour time period.
            Calculate and report the relative response factor and percent
            difference (%D) for all compounds.   The percent difference (%D)
            for each CCC compound must be less  than or equal to 25.0
            percent.  Ensure that the minimum RRF for volatile SPCCs is
            0.300 (0.250 for bromoform).   Additional instructions for
            completing Form VII are found in Exhibit B,  Section III.
PART 3 - METHOD BLANK ANALYSIS

3.     Summary

      The method blank is prepared by adding 100 pL of surrogate to 10 mL
      of methanol.  A volume of the methanol is added to 5 mL of reagent
                                 VOA E-16
                                             Rev. 9/88

-------
      water and carried through the entire analytical scheme.  The method
      blank volume must be approximately equal to the sample volumes being
      processed.

3.1   Method blank analysis must be performed at the following frequency:

      3.1.1    For the  analysis of volatile target compounds, a method blank
               analysis must be performed once for each  twelve hour time
               period.  The twelve hour  period begins with  the injection  of
               BFB.   The method blank must be analyzed after  the calibration
               standard(s).

3.2   It is the Contractor's responsibility to ensure that method
      interferences caused by contaminants in solvents,  reagents,
      glassware, and other sample processing hardware that lead to discrete
      artifacts and/or elevated baselines in gas chromatograms be
      minimized.

      3.2.1    For the  purposes of this  protocol, an acceptable laboratory
               method blank should meet  the criteria of  paragraphs 3.2.1.1
               and 3.2.1.2.

               3.2.1.1    A method blank for volatile analysis must contain
                         less than or equal to five times  (5x) the Contract
                         Required Quantitation Limit (CRQL from Exhibit  C)
                         of methylene chloride, acetone, 2-butanone, and
                         toluene.

               3.2.1.2    For all- other  target compounds not  listed above,
                         the method blank must contain  less  than or equal .
                         to the Contract Required Quantitation Limit of  any
                         single target  analyte.

      3.2.2    If  a laboratory method blank exceeds this criterion, the
               Contractor must consider  the analytical system to be out of
               control.  The source of the contamination must be
               investigated and appropriate corrective measures MUST be
               taken  and documented before further sample analysis proceeds.
               All samples processed with a method blank that is out of
               control  (i.e., contaminated) MUST be reextracted/repurged  and
               reanalyzed at no additional cost to the Agency.  The
               Laboratory Manager, or his designee, must address problems
               and solutions in the Case Narrative (Exhibit B).

3.3   Documentation

      The Contractor  shall report  results  of  method blank  analysis using
      the Organic  Analysis Data Sheet  (Form I)  and the form for  tentatively
      identified compounds (Form  I, TIC).  In addition,  the samples
      associated with each method  blank  must  be  summarized on Form IV
      (Method Blank Summary).  Detailed  instructions  for the  completion of
      these forms  can be found in  Exhibit  B,  Section  III.
                                 VOA E-17                              Rev. 9/88

-------
            3.3.1    The Contractor shall report ALL sample concentration
                     data as UNCORRECTED for blanks.
PART 4 - SURROGATE SPIKE (SSt ANALYSIS

4.    Summary

      Surrogate standard determinations are performed on all samples and
      blanks.  All blanks are fortified with surrogate spiking compounds
      before extraction in order to monitor preparation and analysis of
      samples.

    4.1   Each sample, blank, and control matrix spike must be fortified
          with the surrogate compounds (shown in Table 4.1) prior to
          extraction and purging.  Performance based criteria are generated
          from laboratory results. Therefore, deviations from the spiking
          protocol will not be permitted.


               TABLE 4.1   SURROGATE SPIKING LEVELS AND RECOVERY LIMITS
Compound Concentration*
Toluene-dg
4 - Bromof luor obenzene
1 , 2 - Dichloroe thane - d^
50 ug
50 ug
50 ug
Recovery
Limit (%)**
50
50
50
- 160
- 160
- 160
           *In sample extract at the time of injection (before any optional
             dilutions).
          **These limits are mandatory for method blanks and advisory for
             samples  and  control matrix spikes.  They are not used to
             determine  if a  sample or a control matrix spike should be
             reanalyzed.

4.2   Surrogate spike recovery must be evaluated by determining whether the
      concentration  (measured as percent recovery) falls inside the
      contract required  recovery limits listed in Table 4.1.

4.3   Treatment of surrogate spike recovery information shall be according
      to paragraphs 4.3.1 and 4.3.2.

      4.3.1   Method Blank  Surrogate Spike Recovery

              The laboratory must take actions listed below if recovery  of
              any one  surrogate compound  in  the volatiles  fraction of  the
              method blank  is outside of  the contract required surrogate
              spike  recovery limits  (Table 4.1).
                                 VOA E-18
Rev. 9/88

-------
               4.3.1.1    Check calculations  Co ensure there are no errors;
                          check Internal standard and surrogate spiking
                          solutions for degradation,  contamination, etc;
                          also,  check instrument performance.

               4.3.1.2    Recalculate or re-inject/repurge  the  blank if
                          steps  in 4.3.1.1  indicate  this  may produce
                          compliant surrogate  recoveries.

               4.3.1.3    Re-extract and re-analyze  the blank and any
                          associated samples.

               4.3.1.4    If  the measures listed in  4.3.1.1  thru 4.3.1.3
                          fail  to correct the  problem,  the analytical system
                          must be considered out of  control.  The problem
                          MUST be corrected before continuing.

                          This may mean recalibrating  the instrumentation
                          but it may also mean more  extensive action.   The
                          specific corrective  action is left up  to the  GC/MS
                          operator.   When surrogate  recovery(ies)  in the
                          blank  is outside of  contract required  windows,  all
                          samples associated with that blank MUST be
                          reanalyzed at no additional  cost to the Agency.

      4.3.2    Sample  Surrogate  Spike Recovery..

               When the  recovery of any one surrogate  compound  in the
               volatiles fraction of the sample or control matrix spike is
               outside of the contract surrogate spike recovery  limits^"-
               (Table  4.1), it  is the responsibility of the  Contractor to
               establish that the deviation is not due to laboratory
               problems.   The surrogate spike  recovery windows  for high
               concentration  samples and control matrix spikes  are advisory
               at  this time.

4.4   Documentation

            The Contractor is required to report surrogate recovery data    -
            for the following:

            o  Method Blank Analyses

            o  Sample Analyses

            o  Control Matrix Spike Analyses

            The surrogate spike recovery data are summarized on the
            Surrogate Spike Percent Recovery  Summary  (Form II).  Detailed
            instructions for the completion of Form II are in Exhibit B,
            Section III.
                                 VOA E-19                             Rev.  9/88

-------

PART 5 - CONTROL MATRIX SPIKE ANALYSIS

5.    Summary

      In order to evaluate the efficiency of the analytical methodology,
      the USEPA has developed standards to be used for control matrix spike
      analysis.  These compounds are subject to change depending upon
      availability and suitability for use as control matrix spikes.

5.1   The Contractor shall perform one control matrix spike sample analysis
      for each Case received, or for each 20 single phase units, or each 14
      calendar day period during which single phase units in a Case were
      received (said period beginning with the receipt of the first sample
      in that Sample Delivery Group) whichever is most frequent.

5.2   Use the following compounds to prepare control matrix spiking
      solutions (Exhibit D, Section II, 5.5.1), which require that a
      uniform amount of control matrix spiking solution be added to a
      control matrix aliquot prior to extraction.  The spiking solution
      contains the following compounds:  1,1-dichloroethene,
      trichloroethene, chlorobenzene, toluene, and benzene.

            Analyze the control matrix spike and calculate the individual
            component recoveries using Equation 5.1.
            Control Matrix Spike
            Percent Recovery
CR

SA
100
Equation 5. !•
            where,
            CR  -  Control matrix spike concentration
            SA  -  Concentration of spike added from spiking mix
5.3   Documentation  .

      Matrix spike recovery limits are 60% - 150% for all of the volatile
      control matrix spike compounds.  These limits are for advisory
      purposes only.   (They should not be used to determine reanalysis of
      a control matrix spike.)  When sufficient multi-laboratory data
      become available, standard recovery limits will be calculated.

      The control matrix spike percent recoveries shall be summarized on
      Form III.  These values will be used by EPA to establish performance
      based QC recovery limits.  Complete instructions for the completion
      of Form III can be found in Exhibit B, Section III.

PART 6 - SAMPLE ANALYSIS

6.    Summary

6.1   Samples can be analyzed upon successful completion of the initial
      calibration analysis.  When twelve (12) hours have elapsed since the
      initial calibration was completed, it is necessary to conduct an
                                 VOA E-20
                                Rev.  9/88

-------
instrument tune and continuing calibration analysis.  Any major
system maintenance, such as a source cleaning or installation of a
new column, will necessitate a retune and recalibration  (See Initial
Calibration, Part 3).

6.1.1    Internal  Standards Evaluation

         Internal  standard response and retention times  in all  samples
         must be evaluated immediately after  or  during data
         acquisition.   If the  retention time  for any internal standard
         changes by more than  30  seconds,  the chromatographic system
         must be inspected for malfunctions and  corrections made  as
         required.   If the extracted ion current profile  (EICP)  area
         for  'any internal standard  changes by more  than  a factor  of
         two  (-50%  to  +100%),  from  the latest q"aily (12 hour time
         period) calibration standard,  the mass  spectrometric system
         must be inspected for malfunctions,  and corrections made as
         required.   The extracted ion current profile (EICP) of the
         internal  standards must  be monitored and evaluated for each
         single phase  unit, method  blank and  control matrix. The
         criteria  are  described  in  detail in  the instructions for Form
         VIII,  High Concentration Internal Area  Summary  (See Exhibit
         B).  Breaking off 1 foot of the column  or  cleaning the
         injector  sleeve will  often improve high end sensitivity  for
         the  late  eluting compounds.

         Poor injection technique can also lead  to  variable IS  ratios.
         When corrections are  made,  reanalysis of samples analyzed
         while  the  system was  malfunctioning  is  necessary.

         6.1.1.1    If after reanalysis,  the  EICP areas for  all
                    internal standards are inside the contract  limits
                    (-50%  to +100%),  then the problem with  the  first
                   analysis is considered to have  been within  the
                   control of the  laboratory.   Therefore, only submit
                   data from the analysis with  EICP's with  the
                   contract limits.  This is considered  the initial
                   analysis and must be reported as  such on all data
                   deliverables.

         6.1.1.2    If  the  reanalysis of the sample does  not solve the
                   problem, i.e., the EICP areas are  outside contract
                   limits  for both analyses,  then  submit the EICP
                   data and sample data from both  analyses.
                   Distinguish between the initial analysis and the
                   reanalysis on all data deliverables,   using the
                   sample  suffixes specified in Exhibit  B.  Document
                   in  the  Case Narrative all inspection  and
                   corrective actions taken.

6.1.2    Each analytical run must also be checked for saturation.  The
         level  at which an  individual compound will saturate the
         detection system is a function of the overall system
         sensitivity and the mass spectral characteristics of that

                           VOA E-21                             Rev.  9/88

-------
        compound.  The initial method calibration (Part 2) requires
        that the system should not be saturated for high response
        compounds at 200 ug/L for VGA target compounds.

        6.1.2.1    If any compound in any sample exceeds the initial
                   calibration range, that sample must be diluted,
                   the internal standard concentration readjusted,
                   and the sample re-injected,  as described in
                   specific methodologies in Exhibit D VOA.
                   Secondary ion quantitation is only allowed when
                   there are sample matrix interferences with the
                   primary ion.

        6.1.2.2    If the dilution of the sample extract causes any
                   compound detected in the first analysis to be
                   undetectable in the second analysis, then the
                   results of both analyses shall be reported on
                   separate Forms (I), according to instructions in
                   Exhibit B'.

6.1.3   Qualitative Analysis

        The target compounds listed  in Exhibit C, shall be identified
        by an analyst competent in the interpretation of mass
        spectra, by comparison of the^suspect mass spectrum to the
        mass spectrum of a standard  of the suspected compound.  Two
        criteria must be satisfied to verify the identifications:
        (1) elution of the sample component at the same GC relative
        retention time as .the standard component, and (2)
        correspondence of the sample component and standard component
        mass spectra  (Exhibit D).

        6.1.3.1    For establishing correspondence of the GC relative
                   retention time (RRT), the sample component RRT
                   must compare within ±0.06 RRT units of the RRT of
                   the standard component.  For reference, the
                   standard must be run on the same shift as the
                   sample.

        6.1.3.2    For comparison of standard and sample component
                   mass spectra, mass spectra obtained on the
                   Contractor's GC/MS are required.   The BFB tuning
                   requirements listed in Part 1 of this Section must
                   be met on that same GC/MS.

        6.1.3.3    The requirements for qualitative verification by
                   comparison of mass spectra are as follows:

                   6.1.3.3.1    All ions present in the standard mass
                                spectra at a relative intensity
                                greater than 10% (most abundant ion
                                in the spectrum equals 100%) must be
                                present in the  sample spectrum.
                           VOA E-22
Rev. 9/88

-------
                   6.1.3.3.2    The relative Intensities of ions
                                specified in the above paragraph must
                                agree within ±20% between the
                                standard and sample spectra.

                   6.1.3.3.3    Ions greater than 10% in the sample
                                spectrum but not present in the
                                standard spectrum must be considered
                                and accounted for by the analyst
                                making the comparison.   When GC/MS
                                computer data processing programs are
                                used to obtain the sample component
                                spectrum,  both the processed and the
                                raw spectra must be evaluated.   The
                                verification process should favor
                                false positives.

        6.1.3.4    If a compound cannot be verified by all of the
                   criteria in 6.1.3.3, but in the technical
                   judgement of the mass spectral interpretation
                   specialist the identification is correct, the
                   Contractor shall report the identification and
                   proceed with the quantitation.

        6.1.3.5    A library search shall be executed for
                   nonsurrogate and non-target sample components for
                   the purpose of tentative identification.  For this
                   purpose, the 1985 or most recent available version
                   of the National Bureau of Standards Mass Spectral
                   Library, containing 42,261 spectra should be used.
6.1.4   Quantitation
        6.1.4.1    Target compounds identified shall be quantitated      [
                   by the internal standard method.  The internal
                   standards used shall be the ones assigned in Table
                   2.1 of this Section.  The EICP area of
                   characteristic ions of target analytes are used
                   (Exhibit D).

        6.1.4.2    An estimated concentration for non-target
                   compounds tentatively identified shall be
                   quantitated by the internal standard method.  For
                   quantification, the nearest internal standard free
                   of interferences must be used.

        6.1.4.3    Calculate surrogate standard recovery for all
                   surrogate compounds, on all single phase units,
                   method blanks, and control matrix spikes.  If
                   recovery is within contractual  or advisory limits,
                   report on Form II.   If recovery is outside
                   contractual limits for the method blank, take
                   specific steps listed in Section 4.3, Surrogate
                   Spike Recoveries.

                           VQA E-23                             Rev. 9/88

-------
        6.1.4.4    Calculate control matrix spike present recovery
                   for all spiked compounds.  Report results on Form
                   III.

6.1.5   Reporting and Deliverables

        Refer to Exhibit B of this document for specific details on
        contract deliverables and reporting formats.  Exhibit B
        contains specific instructions for completing all required
        forms, as well as a detailed itemization of reporting and
        deliverables requirements.
                           VOA E-24
Rev. 9/88

-------
          SECTION III
EXTRACTABLES QA/QC REQUIREMENTS
          EXT  E-25
Rev. 9/88

-------
                              TABLE OF CONTENTS
Part 1 - Tuning and GC/MS Mass Calibration
Part 2 - Calibration of the GC/ MS System
Part 3 - Method Blank Analysis
Part 4 - Surrogate Spike (SS) Analysis
Part 5 - Control Matrix Spike Analysis
Fart 6 - Sample Analysis
PAGE NUMBER




  EXT  E-27






  EXT  E-28







  EXT  E-35






  EXT  E-36






  EXT  E-38






  EXT  E-40
                                 EXT E-26
             Rev. 9/88

-------
This Section outlines the minimum Quality Control (QC) operations necessary
to satisfy the analytical requirements.associated with the determination of
high concentration extractable target compounds in waste samples.  These QC
operations are as follows:

     o   Documentation of  GC/MS Mass  Calibration and Abundance Pattern
     o   Documentation of  GC/MS Response Factor Stability
     o   Internal  Standard Response and Retention Time Monitoring
     o   Method Blank Analysis
     o   Surrogate Spike Response Monitoring
     o   Control Matrix Spike Analysis


Fart 1 - Tuning and GC/MS MassCalibration

1.    Summary

      Prior to initiating any on going data collection,  it is necessary to
      establish that a given GC/MS meets the standard mass spectral abundance
      criteria.  This is accomplished through the analysis of
      Decafluorotriphenylphosphine (DFTPP).  The ion abundance criteria for
      each calibration compound MUST be met before any single phase units,
      method blanks, or control matrix spikes can be analyzed.

1.1   Decafluorotriphenylphosphine (DFTPP)

      1.1.1   Each  GC/MS  system used for  the analysis of semi-volatile or
              pesticide target compounds must be hardware tuned  to meet the
              abundance criteria listed in Table E-l for a 50 ng injection of
              decafluorotriphenylphosphine (DFTPP).  DFTPP may be analyzed
              separately  or as part  of the calibration standard.  The
              criteria must be demonstrated daily or for each twelve (12)
              hour  period, whichever is more frequent.  DFTPP must be
              injected to meet this  criterion.  Background subtraction should
              be  straightforward and designed only to eliminate  column bleed
              or  instrument background ions.  Background subtraction actions
              resulting in spectral  distortions for the sole purpose of
              meeting the contract specifications are contrary to the
              objectives of Quality  Assurance and are unacceptable.

      NOTE:    All instrument conditions must be identical to those in sample
              analyses, except that  a different (faster) temperature program
              may be used.

      1.1.2    Whenever the Contractor takes corrective action which may
              change or affect the tuning criteria of DFTPP (e.g., ion source
              cleaning or repair, etc.), the tune must be verified
              irrespective of the twelve (12) hour tuning requirements.

      1.1.3    Any samples analyzed when tuning criteria have not been met
              will require reanalysis at no additional cost to the Agency.
              The twelve (12.0) hour time period for GC/MS system tuning and
              standards calibration  (initial or continuing calibration

                                 EXT  E-27                           Rev. 9/88

-------
              criteria) begins at the moment of injection of the DFTPP
              analysis that the laboratory submits as documentation of
              compliant tune.  The time period ends after twelve (12) hours
              has elapsed according to the GC/MS system clock.

1.2   Documentation

      Documentation of the calibration must be provided in the form of a bar
      graph plot and as a mass listing.

      1.2.1   The Contractor shall complete a Form V (High Concentration
              Extractable GC/MS Tuning and Mass Calibration) each time an
              analytical system is tuned.  In addition, all single phase
              units, standards, method blanks, and control matrix spikes
              analyzed during a particular tune must be summarized on the
              bottom of the appropriate Form V.  Detailed instructions for
              the completion of Form V are found in Exhibit B, Section III.
    TABLE £-1.   DFTPP Key Ions and Ion Abundance Criteria

                                   Ion Abundance Criteria
«•••••••
51 .
68
69 '
70
127
197
198
199
275
365
441
442
443
Part 2 - Calibration of
30.0 to 60.0 percent of mass 198
Less than 2.0 percent of mass 69
Mass 69 relative abundance
Less than 2.0 percent of mass 69
40.0 to 60.0 percent of mass 198
Less than 1.0 percent of mass 198






Base peak, 100 percent relative abundance
5.0 to 9.0 percent of mass 198
10.0 to 30.0 percent of mass 198
Greater than 1.00 percent of mass
Present but less than mass 443
Greater than 40.0 percent of mass
17.0 to 23.0 percent of mass 442
the GC/MS Svstem


198

198


2 . Summary
      Prior to the analysis of samples,  method blanks,  or control matrix
      spikes, and after tuning criteria have been met,  the GC/MS system must
      be initially calibrated at a minimum of three concentrations to
                                 EXT E-28                          Rev. 9/88

-------
      determine the linearity of response utilizing target compound
      standards.  Once the system has been calibrated, the calibration must
      be verified each twelve (12) hour time period for each GC/MS system.

2.1   Prepare calibration standards as described in Exhibit D, Section 4, to
      yield the following specific concentrations:

      2.1.1   Extractable Target Compounds

              Initial calibration of all extractable  target compounds except
              PCBs are required at 50, 80 and 160 total nanograms.  Initial
              calibration of PCS homolog:  one solution containing mono-
              through heptachlorobiphenyl are required at 10, 30 and 50  total
              nanograms.  Octa-, nona- and decachlorobiphenyl are required at
              20, 60 and 100 total nanograms (due to  their low responses).

2.2   The USEPA plans to develop  performance based criteria for response
      factor data acquired during this program.   To accomplish this goal, the
      Agency has specified both the concentration levels  for initial
      calibration and has also specified the internal standard to be used on
      a compound by compound basis for quantitation.   Establishment of
      standard calibration procedures is necessary and deviations by
      contractor laboratories will not be allowed.

2.3   Analysis of Calibration Standard

      2.3.1   Analyze each calibration standard and tabulate the area of  the
              primary characteristic ion (Exhibit D,  Table 1) against
              concentration for each compound including all contract required
              surrogate compounds.  The relative retention times of each  •
              compound in each calibration run should agree within ±0.05  RRT.
              Late eluting compounds usually will have much better agreement.
              Using Table E-3, calculate the relative response factors  (RRF)
              for each target compound at each concentration level using
              Equation 3.1.

      Equation 3.1

            RRF •
      Where:

      AX - Area of the characteristic ion for the compound to be
           measured.

      A^s - Area of the characteristic ion for the specified internal
           standards from Table E-3.

      Cis ~ Concentrati°n of the internal standard (ng/uL).

      Cx =• Concentration of the  compound to  be measured (ng/uL).
                                 EXT  E-29
Rev. 9/88

-------
      2.3.2   Using the relative response factors (RRF) from the initial
              calibration, calculate the percent relative standard deviations
              (% RSD) for compounds labeled as Calibration Check Compounds
              using Equation 3.2.

              Equation 3.2

              % RSD -  SD  x 100
                       x

              Where:

              RSD - Relative Standard Deviation.

              SD  - Standard Deviation of 3 Initial Relative
                    Response Factors (Per Compound).

              Where:
                        n-1
              x - Mean of 3 Initial Relative Response Factors (Per Compound).

              The % RSD for each individual Calibration Check Compound must
              be less than 30.0 percent.  This criteria must be met for the
              initial calibration to be valid.

2.4   A system performance check must be performed to ensure minimum average
      relative response factors are met before the calibration curve is used.

      2.4.1   For extractables, the System Performance Check Compounds
              (SPCC's) are:

              N-Nitroso-di-n-propylamine, Hexachlorocyclopentadiene, 2,4-
              Dinitrophenol and 4-Nitrophenol.  The minimum acceptable
              average relative response factor (RRF) for these compounds is
              0.050.  These compounds  (SPCC's) typically have very low RRF's
              (0.1-0.2) and tend to decrease in response as the
              chromatographic  system or the standard material begin to
              deteriorate.  These compounds are usually the first to show
              poor performance.  Therefore, they  must meet the minimum
              requirement when the system is calibrated.

      2.4.2   The initial calibration  is valid only after both the % RSD for
              CCC compounds and the minimum RF for SPCC have been met.  Only
              after both these criteria are met can sample analysis begin.
                                 EXT E-30
Rev. 9/88

-------
2.5   Documentation

      Once the initial calibration is validated, calculate and report the
      average relative response factor (RRF) and percent relative standard
      deviation (% RSD) for all target compounds.  The laboratory is required
      to submit a Form VI (Initial Calibration Data) for each instrument used
      to analyze samples under this protocol.   Detailed instructions for
      completion of Form VI are found in Exhibit B,  Section III.

2.6   Continuing Calibration

      The analysis of a calibration standard(s) containing all required
      target compounds, including all required surrogates, must be performed
      each twelve (12) hours during analysis.   Compare the response factor
      data from the standards each twelve hours with the average  relative
      response factor from the initial calibration for a specific instrument.
      A system performance check must be made  each twelve (12) hours.  If the
      SFCC criteria are met, a comparison of response factors is  made  for all
      compounds.   This is the same check that  is applied during the initial
      calibration (Form VI).  If the minimum response factors are not  met,
      the system must be evaluated and corrective action must be  taken before
      sample analysis begins.

      2.6.1   Some possible problems are standard mixture degradation,
              injection port inlet contamination, contamination at the front-
              end of  the analytical column, and active sites  in the column or
              chromatogfaphy system. This check must be met before analysis
              begins.  The minimum relative response factor  (RRF) for
              extractable System Performance Check Compounds  (SPCC) is 0.050.

      2.6.2   Calibration Check Compounds (CCC)

              After the system performance check is met, Calibration Check
              Compounds listed in Table E-2 are used to check the validity of
              the initial calibration.  Calculate the percent difference
              using Equation 3.3.

              Equation 3.3
                  % Difference - RRFj -  RRFC
                                             x 100
                                    RRF,
              Where:
              RRFj - Average Relative Response Factor from Initial
              Calibration.

              RRFC - Relative Response Factor from Current Verification Check
              S tandard.
                                 EXT  E-31
Rev. 9/88

-------
              2.6.2.1    If the percent difference for any compound is
                         greater than 20.0%, the laboratory should consider
                         this a warning limit.  If the percent difference for
                         each CCC is less than 25.0%, the initial calibration
                         is assumed to be valid.  If the criteria are not met
                         (greater than 25.0% difference), for any one
                         calibration check compound, corrective action MUST
                         be taken.  Problems similar to those listed under
                         SPCC could affect this criteria.  If no source of
                         the problem can be determined after corrective
                         action has been taken, a new initial five point
                         calibration MUST be generated.  This criteria MUST
                         be met before sample analysis begins.
TABLE E-2.  Extractable Calibration Check Compounds
     Acenaphthene
     1,4-Dichlorobenzene
     Hexachlorobutadiene
     N-Nitroso-di-n-phenylamine
     Di-n-octyIphthalate
     Fluoranthene
     Benz o(a}pyrene
             4-Chloro- 3-Methylphenol
             2,4-Dichlorophenol
             2-Nitrophenol
             Phenol
             Pentachlorophenol
             2,4,6-Trichlorophenol
                                                   I
      2.6.3   Concentration Levels for Continuing Calibration Check

              The USEPA plans to evaluate the long term stability of response
              factors during this program. Standardization among contract
              laboratories is necessary to reach these long term goals.
              Along with contract specified concentrations for initial
              calibration, the USEPA  is requiring specific concentrations for
              each calibration standard(s).
               2.6.3.1
The concentration for each extractable target
compound except PCBs in the continuing calibration
standard(s) is 80 total nanograms.  For the
continuing calibration of the PCBs, one solution
containing mono-through heptachlorobiphenyl is
required at 30 total nanograms, octa-, nano-, and
decachlorobiphenyl are required at 60 total
nanograms for the continuing calibration solution.
2.7   Documentation
               The  laboratory is  required  to  complete and  submit a  Form VII
               for  each GC/MS system utilized for  each  twelve  (12)  hour time
               period.   Calculate and report  the response  factor and  percent
               difference  (%  D) for  all  compounds.   Ensure the minimum RF  for
               SPCC's  is 0.050.   The percent  difference (% D) for each CCC
                                 EXT E-32
                                          Rev.  9/88

-------
compound must be less than 25.0 percent.   Additional
instructions for completing Form VII are  found in Exhibit B,
Deliverables, Section III.
                  EXT E-33
Rev. 9/88

-------
TABLE E-3.  Standards With Corresponding Extractable Target Compounds
Analytes Assigned For Quantitation
1,4-Dichlorobenzene-d.
Naphthalene-do
Ac enaphthene - d-[ n
Phenol

bis(2-Chloroethyl)ether
2-Chlorophenol
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Benzyl alcohol
1,2-Dichlorobenzene
2-Me thyIpheno1
bis(2-Chloroisopropyl)ether
4 -Methylphenol
N-Nitroso-di-n-propylamine
Hexachloroethane
2-Fluorophenol  (surr)
Phenol"-d5  (surr)
Nitrobenzene

Isophorone
2-Nitrophenol
2,4-Dimethylphenol
Benzole acid
bis(2-Chloroethoxy)raethane
2,4-Dichlorophenol
1,2,4-Trichlorobenzene
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
4-Chloro-3-methylphenol
2-Me thylnaphthalene
Nitrobenzene-d5 (s.urr)
 sxirr —  surrogate compound
Hexachlorocyclopenta-
diene
2,4,6-Trichlorophenol
2,4,5-Trichlorophenol
2-Chloronaphthalene
2-Nitroaniline
Dimethyl phthalate
Acenaphthylene
3-Nitroaniline
Acenaphthene
2,4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Diethyl phthalate
4-Chlorophenyl-phenyl
ether
Fluorene
4-Nitroaniline
2-Fluorob ipheny1
(surr)
2,4,6-Tribromophenol
 (surr)
                                  EXT E-34
                                       Rev. 9/88

-------
TABLE E-3.   (Continued)
Phenanthrene-
 Chrysene-di2
Perylene-d]^
4,6-Dinitro-2-methylphenol
N-nitrosodiphenylamine
1,2-Diphenylhydrazine
4-Bromophenyl phenyl ether
Hexachlorobenzene
Pentachloropheno1
Phenanthrene
Anthracene
Di-n-butyl phthalate
Fluoranthene
alpha-BHC
beta-BHC
gamma-BHC
delta-BHC
Heptachlor
Aldrin
Heptaclor epoxide
Endrin ketone
 Pyrene
 Butylbenzyl phthalate
 3,3'-Dichlorobenztdine
 Benzo(a)anthracene
 bis(2 -ethyIhexyl)phthalate
 Chrysene
 Terphenyl-dl4 (surr)
 gamma-Chlordane
 Endosulfan I
 Alpha-Chlordane
 Dieldrin
 4,4'-DDE
 Endrin
 Endosulfan II
 4,4'-ODD
 Endosulfan sulfate
 4,4'-DDT
 Methoxychlor
 Konochlorob ipheny1
 Dichlorobiphenyl
 Trichlorobiphenyl
 Tetrachlorobiphenyl
 Pentachlorobipheny1
 Hexachlorobipheny1
 Heptachlorobiphenyl
. Octachlorobipheny1
 Nonachlorobiphenyl
 Decachlorobipheny1
Di-n-octyl phthalate
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benz o(a)pyrene
Indeno(l,2,3-cd)
  Pyrene
Dibenz(a,h)anthracene
Benzo(g,h,i)perylene
      surr - surrogate compound
                                 EXT  E-35
                                       Rev. 4/89

-------
Part 3 - Method Blank Analysis Summary

3.     Summary

      A method blank is an aliquot of the reagent(s) that is carried through the
      entire analytical scheme (extraction, concentration, and analysis).   The
      method blank weight must be approximately equal to the sample weights being
      processed.

3.1   Method blank analysis must be performed at the following frequency:

      3.1.1    For  the  analysis of extractable  target compounds,  a  method  blank
               analysis must be performed  for each case  received, or  for each  20
               single phase units, or whenever  single phase units are extracted,
               whichever  is most  frequent.  The method blank associated  with a
               specific set or group of  single phase units must be  analyzed on each
               GC/MS or GC system used to  analyze that specific group or set.

      3.1.2    It is the  laboratory's responsibility to  ensure that method
               interferences caused by contaminants  in solvents,  reagents,
               glassware, and other sample processing hardware that lead.to
               discrete artifacts and/or elevated baselines in gas  chromatograms be
               minimized.

               3.1.2.1    For the purposes  of this protocol, an acceptable
                         laboratory method blank must meet the criteria of
                         paragraphs 3.1.2.1.1 and 3.1.2.1.2.

                         3.1.2.1.1'    A method blank for extractable analysis  must
                                      contain less  than two times (2X)  the
                                      Contract Required Quantitation Limit (CRQL
                                      from Exhibit  C)  of phthalate esters  in the
                                      TCL.

                         3.1.2.1.2    For all other extractable  target compounds
                                      not listed above,  the method blank must
                                      contain less  than or equal to the Contract
                                      Required Quantitation Limit (CRQL) of any
                                      target analyte.

      3.2.2    If a laboratory method blank exceeds criteria, the Contractor must
               consider the analytical system out of control.  The  source  of the
               contamination must be investigated and appropriate corrective
               measures must be taken before further sample analysis proceeds.  All
               single phase units processed with a method blank that  is  out of
               control  (i.e., contaminated) must be reextracted and reanalyzed at
               no additional cost to the Agency.  The Laboratory Manager,  or his
               designate, must address problems and  solutions in  the Case  Narrative
               (Exhibit B).
                                   EXT E-36
Rev. 9/88

-------
3.3   Documentation

      Results of extractable method blank analysis shall be reported using the
      High Concentration Extractable Analysis Data Sheet (Form I), and the
      tentatively identified compounds (Form I, TIC).  In addition, the single
      phase units associated with each method blank must be summarized on Form IV
      (High Concentration Extractable Method Blank Summary).   Specific
     . instructions for the completion of these forms can be found in Exhibit B
      (Reporting and Deliverables),  Section III.

      3.3.1   The Contractor will report ALL sample  concentration data as
              UNCORRECTED  for  blanks.  It shall be the responsibility of the  EPA
              evaluator, and/or data  auditor,  to  correct analyte  concentrations
              for concentrations detected in the  method blank(s).   It is the
              Contractor's responsibility to ensure  the proper number of method
              blanks  are analyzed and the data properly reported.

       - Surrogate (SS) Analysis

4.    Summary

      Surrogate standard determinations are performed on all single phase units,
      method blanks, and control matrix spikes.  ALL single phase units,  method
      blanks, and control matrix spikes are fortified with surrogate spiking
      compounds before extraction in order to monitor preparation and analysis of
      samples.

4.1   Each single phase unit (including control matrix spike and method blank) is
      spiked with surrogate compounds prior to extraction.   The surrogate spiking
      compounds shown in Table E-4 are used to fortify each single phase unit with
      the proper concentrations.  Performance based criteria are generated from
      laboratory results.   Therefore deviations from spiking protocol will not be
      permitted.
              TABLE E-4.Extractable Surrogate Spiking Compounds
            Compound

            Nitrobenzene-dij
            2 -Fluorobipheny1
            p-Terphenyl-d^
            Phenol-de
            2-Fluorophenol
            2,4,6-Tribromophenol
   Amount in Sample Extract
(before any optional dilutions')

             50 ug
             50 ug
             50 ug
            100 ug
            100 ug
            100 ug
4.2   Surrogate recovery must be evaluated for acceptance by determining whether
      the concentration (measured as percent recovery) falls inside the advisory
      recovery limits listed in Table E-5.
                                   EXT E-37
                   Rev. 9/88

-------
4.3   Surrogate'spike recovery information is treated according to paragraphs
      4.3.1 through 4.3.1.2.

      4.3.1   Method Blank  Surrogate Recovery

              When the  surrogate  recovery  of any  one  extractable  surrogate
              compound  is outside of the contract required surrogate  recovery
              limits (listed in Table  E-5)  for  a  method blank,  the  laboratory must
              take the  following  actions:

              4.3.1.1   Check calculations  to  assure there are no  errors.   Check
                         internal standard and  surrogate  spiking  solutions  for
                         degradation,  contamination,  etc.   Also,  check instrument
                         performance.

              4.3.1.2   Recalculate and/or reinject  the  extract  if steps  in
                         4.3.1.1  reveal the  cause of  the  non-compliant surrogate
                         recoveries.

              4.3.1.3   If the measures listed above fail to  correct the  problem,
                         the analytical system  must be  considered out of control.
                         The problem MUST  be corrected  before  continuing.   This
                         may mean recalibration of the  instrumentation but  it may
                         also mean more extensive action.   The specific corrective.
                         action is left up to the GC/MS operator.   The method
                         blank and all associated single  phase units,  including
                         any control matrix spikes must be re-extracted and re-
                         analyzed at no additional cost to the Agency.
               TABLE E-5. Surrogate Recovery Limits*

                      Surrogate  Compound

                      Nitrobenzene -de
                      2 - Fluorob iphenyl
                      p - Terpheny 1 - d^
                      Phenol -dg
                      2-Fluorophenol
                      2,4, 6 -Tribromophenol
              %  Recovery

                20-140
                20-140
                20-150
                20-140
                20-140
                10-140
               *  Mandatory for method blanks .
               control matrix spikes.
Advisory for single phase units and
                                    EXT  £-38
                      Rev. 9/88

-------
      4.3.2    Sample  Surrogate  Recovery

               When the  surrogate  recovery  of  any  one  extractable  surrogate
               compound  is  outside of the advisory recovery limits (listed in
               Table E-5)  for any  single phase unit (including the control matrix
               spike(s)),  the laboratory must  take the following actions:

               4.3.2.1   Check  to be sure  there are no errors  in  calculations,
                         surrogate solutions  and  internal  standards.  Also,  check
                         instrument  performance.

               4.3.2.2   Recalculate the data and/or  reanalyze  the extract if any
                         of  the above checks  reveal a problem.

               4.3.2.3   If  surrogates  cannot be  detected  due  to  dilution factors,
                         this requirement  (4.3.2.1) need not be met,  report the
                         value(s) as (DL)  as  described in  Exhibit B.

4.4   Documentation
                                  *
      The Contractor  is required  to  report surrogate recovery data for the
      following:

            o   Method Blanks

    ":      o   Control  Matrix Spikes

            o   ALL Single  Phase Units

      The surrogate spike re~cbvery data is  summarized on the  High Concentration
      Extractable Surrogate Spike Percent Recovery Summary (Form II).  Complete
      instructions for the completion of Form II can be found in Exhibit B,
      Section III.

Part 5 - Control Matrix Spike

5.    Summary

      In order to evaluate the efficiency of the analytical methodology, the USEPA
      has developed the  standard mixtures listed in Table E-6 to be used for
      control matrix spike analysis.   These compounds are subject to change
      depending upon availability and suitability for use as  control matrix
      spikes.  The control matrix spike analysis is not to be misinterpreted as a
      matrix spike or matrix spike duplicate analysis.

5.1   Frequency of Analysis

      The Contractor shall perform one contol matrix spike sample analysis for
      each Case received,  or for each 20 single phase units,  or each 14 calendar
      day period during which single  phase  units in a Case were received (said
      period beginning with the  receipt of  the first sample in that Sample
      Delivery Group)  whichever  is most frequent.
                                   EXT E-39                          Rev.  9/88

-------
5.2   Use the compounds listed in Table E-6 to prepare control matrix spiking
      solutions according to protocols described in  Exhibit D, Section II, 4.8.
      The analytical protocols in Exhibit D,  require that a uniform amount of
      control matrix spiking solution be added to the control matrix prior to
      extraction.

5.3   Individual component recoveries of the control matrix spike are calculated
      using Equation 5.1.
      Matrix Spike Percent Recovery -  SSR - SR x
                                         SA
      Where:

            SSR  - Spike Sample Results
            SR   - Sample Result             '
            SA   - Spike Added from Spiking Mix
                      Equation  5.1
                TABLE E-6.Control Matrix Spiking Solutions*
            BaseyNeutrals

            1,4-Dichlorobenzene
            N-Nitroso-di-n-propylamine
            •1,2,4-Trichlorobenzene
            Acenaphthene
            2,4-Dinitrotoluene
            Pyrene
Acids

Phenol
2 -Chioropheno1
4-Chloro-3-methylphenol
4-Nitrophenol
Penfachlorophenol
Pesticides

Heptachlor
Dieldr-in
            *Base/Neutrals and Acids are combined in one solution.
            are prepared in a separate solution.
                            The pesticides
5.4   Documentation

      The control matrix spike percent recoveries shall be  summarized on Form
      III.  These values will be used by EPA to periodically update existing
      performance based QC recovery limits.

      See Exhibit B, Deliverables,  Section III, for complete instructions on the
      completion of Form III.
                                   EXT E-40
                              Rev. 9/88

-------
      TABLE E-7.Control Matrix Spike Recovery Limits*

            Compound Class     Matrix Spike Compound
            BN
            BN
            BN
            BN
            BN
            BN

            Acid
            Acid
            Acid
            Acid
            Acid

            Pesticide
            Pesticide
N-Nitroso-di-n-propylamine
1,4-Dichlorobenzene
1,2,4-Trichlorobenzene
Acenaphthene
2,4-Dinitrotoluene
Pyrene

Phenol
2-Chlorophenol
4-Chloro-3-methylphenol
4-Nitrophenol
Pentachlorophenol

Heptachlor
Dieldrin
% Recovery*

30-140
30-140
30-140
30-140
30-140
30-140

10-120
10-120
10-120
10-120
10-120

30-140
30-140
      *  These limits are for advisory purposes only.   They are not to be used to
      determine if a sample should be reanalyzed.   When sufficient data are
      available, standard limits will be calculated.

       - Quality Control

6'.    Summary

6.1   Samples can be analyzed upon successful completion of the initial QC
      activities.  When twelve (12) hours have elapsed since the initial QC was
      completed, it is necessary to conduct an instrument tune and calibration
      check analysis (described in Part 1 of this  Exhibit).  Any major system
      maintenance, such as a source cleaning or installation of a new column, will
      necessitate a retune and recalibration (See  Initial Calibration, Section
      III,  Part 2).

      6.1.1    Internal  Standards Evaluation

               Internal  standard response and. retention times  in all samples  must
               be  evaluated immediately after or during data acquisition.   If the
               retention time  for any  internal  standard changes by more than  30
               seconds,  the chromatographic system must be  inspected for
               malfunctions and corrections made as required.  If the  extracted ion
               current profile (EICP)  area for  any internal standard changes  by
               more than a  factor of two  (-50%  to +100%), from the latest daily (12
               hour time  period) calibration standard, the mass spectrometric
               system must be  inspected for malfunction and corrections made  as
               appropriate.  The extracted ion  current profile (EICP)  of the
               internal  standards must be monitored and evaluated for  each  single
               phase unit, method blank, and control matrix.  The criteria  are
               described  in detail in  the instructions for Form VIII,  High
               Concentration Internal Area Summary (See Exhibit B).  Breaking off 1
               foot of the column or cleaning the injector sleeve will often
               improve high end sensitivity for the late eluting compounds.
                                   EXT E-41
                                      Rev. 9/88

-------
        Poor injection technique can also lead to variable IS ratios.  When
        corrections are made, reanalysis of samples analyzed while the
        system was malfunctioning is necessary.

        6.1.1.1    If, after reanalysis, the EICP areas for all internal
                   standards are inside the contract limits (-50% to +100%),
                   then the problem with the first analysis is considered to
                   have been within the control of the laboratory.
                   Therefore, only submit data from the analysis with EICP's
                   within the contract limits.  This is considered the
                   initial analysis and must be reported as such on all data
                   deliverables.

        6.1.1.2    If the reanalysis of the sample does not solve the
                   problem, i.e., the EICP areas are outside contract limits
                   for both analyses,  then submit the EICP data and sample
                   data from both analyses.   Distinguish between the initial
                   analysis and the reanalysis on all data deliverables,
                   using the sample suffixes specified in Exhibit B.
                   Document in the Case Narrative all inspection and
                   corrective actions taken.

6.1.2   Each analytical run must also be checked for saturation.  The level
        at which an individual compound will saturate the detection system
        is a function of the overall system sensitivity and the mass
        spectral characteristics of that compound.  The initial method
        calibration (Part 2) requires that the system should not be
        saturated for high response compounds at 160 nanograms.

        6.1.2.1    If any compound in any sample exceeds the initial
                   calibration range,  that sample must be diluted, the
                   internal standard concentration re-adjusted, and the
                   sample re-injected, as described in Exhibit D, Section
                   III, 5.2.  Secondary ion quantitation is only allowed
                   when there are sample matrix interferences with the
                   primary ion.

        6.1.2.2    If the dilution of the sample extract causes any compound
                   detected.in the first analysis to be undetectable in the
                   second analysis, then the results of both analyses shall
                   be reported on Separate Forms I, according to
                   instructions in Exhibit B.

6.1.3   Qualitative Analysis

        The target compounds listed in Exhibit C, shall be identified by an
        analyst competent  in the interpretation of mass spectra, by
        comparison of  the  suspect mass spectrum to the mass spectrum of a
        standard of the suspected compound.  Two criteria must be satisfied
        to verify the  identifications:  (1) elution of the sample component
        at the same GC relative retention time as the standard component,
        and (2) correspondence of the sample component and standard
        component mass spectra (Exhibit D).
                             EXT E-42
Rev. 9/88

-------
         6.1.3.1    For establishing correspondence of the GC relative
                    retention time  (RRT), the sample component RRT must
                    compare within  +0.05 RRT units of the RRT of the standard
                    component.  For reference, the standard must be run on
                    the same shift  as the sample.

         6'. 1.3.2    For comparison  of standard and sample component mass
                    spectra, mass spectra obtained on the Contractor's GC/MS
                    are required.   The DFTPP tuning requirements listed in
                    Exhibit E, Part 1 must be met on the Contractor's same
                    GC/MS.

         6.1.3.3    The requirements for qualitative verification by
                    comparison of mass spectra are as follows:

                    6.1.3.3.1    All ions present in the standard mass
                                spectra at a relative intensity greater than
                                10% (most abundant ion in the spectrum
                                equals 100%) must be present in the same
                                spectrum.

                    6.1.3.3.2    The relative intensities of ions specified
                                in the above paragraph must agree within
                                ±20% between the standard and sample
                                spectra.

                    6.1.3,3.3    Ions greater than 10% in the sample spectrum
                                but not present in the standard spectrum
                                must be considered and accounted for by the
                                analyst making the comparison.   When GC/MS
                                computer data processing programs are used
                                to obtain the sample component spectrum,
                                both the processed and the raw spectra must
                                be evaluated.   The verification process
                                should favor false positives (Exhibit D).

         6.1.3.4    A library search shall be executed for non-target sample
                    components for  the purpose of tentative identification.
                    For this purpose,  the most recent available version of
                    the EPA/NBS Mass Spectral Library should be used.
6.1.4   Quantitation
        6.1.4.1    Target components identified shall be quantitated by the
                   internal standard method.  The internal standard used
                   shall be the ones assigned in Table E-3, Exhibit E.  The
                   EICP area of characteristic ions of target analytes are
                   used (Exhibit D tables D-l, D-5).

        6.1.4.2    An estimated concentration for non-target components '
                   tentatively identified shall be quantitated by the
                   internal standard method.  For quantification, the
                   nearest internal standard free of interferences must be
                   used.

                             EXT E-43                          Rev. 9/88

-------
I
                          6.1.4.3    Calculate surrogate standard recovery for all surrogate *
                                     compounds, on all single phase units, method blanks, and
                                     control matrix spikes.  If recovery is within contractual
                                     or advisory limits, report on Form II (See Exhibit B).
                                     If recovery is outside contractual limits for the method
                                     blank, take specific steps listed in Exhibit E, Surrogate
                                     Spike Recoveries.

                          6.1.4.4    Calculate control matrix spike percent recovery for all
                                     spiked compounds.  Report results on Form III.

                  6.1.5   Reporting  and Deliverables

                          Refer to Exhibit B  of  this document for specific details  on  contract
                          deliverables and reporting formats.  Exhibit B  contains specific
                          instructions for completing  all  required Forms., as  well as a
                          detailed itemization of reporting and deliverables  requirements.
                                               EXT E-44
Rev. 9/88

-------
              EXHIBIT E
AROCLORS/TOXAPHENE QA/QC REQUIREMENTS
             ARO  E-45
Rev. 9/88

-------
1.
Summary
2.
This section outlines the minimum quality control (QC) operations
necessary to satisfy the analytical requirements associated with the
determination of Aroclors and Toxaphene in high concentration samples
using the analytical method described in Exhibit D.  The QC operations
described below apply to analyses performed on both GC columns of all
instruments.

Initial QC requirements:

  o  Determination of potential quantitation peaks for each analyte and
     determination of their Absolute Retention Times (RT) and
     calibration factors.

Periodic PC requirements:

  o  Continued monitoring of RTs and calibration factors.
  o  Instrument blank analysis.
  o  Method blank analysis.
  o  Control matrix spike analysis.
Additional PC requirements:

 o   Surrogate recoveries are reported for all samples, blanks,  and
     control matrix spike analyses.

 o   Retention time shifts of the surrogates are reported for the
     analyses of all standards, samples, control matrix spikes and
     blanks.

These QC operations are designed to facilitate comparison of analytical
data from different laboratories.  These requirements do not release
the analytical laboratory from maintaining their own checks on the
performance of their instruments.  These checks may include, but are
not limited to, determining detector standing current, monitoring the
number of theoretical plates per column and establishing the limits of
quantitation for each of the analytes.

DEFINITIONS
 2.1   Instrument Blank

      The instrument blank is a solution containing the surrogates
      Tetrachloro-meta-xylene and Decachlorobiphenyl (20 ng/mL) in hexane.
      This solution must be injected no less than once every 12 hours on each
      GC column used in order to demonstrate that none of the Aroclors or
      Toxaphene listed in the Exhibit C are detected at > 0.5 the contract
      required quantitation limit (CRQL).

 2.2   Method Blank

      A method blank is 500 mg of corn oil spiked with the surrogates (20
      ng/mL) that is carried through the entire analytical scheme given in
      Exhibit D.  An acceptable method blank is required for each full or
                                 ARO E-46
                                                                 Rev.  9/88

-------
      partial set of 20 samples analyzed in a sample delivery group (SDG).
      An acceptable method blank has none of the Aroclors or Toxaphene listed
      in Exhibit C at > CRQL.

2.3   Control Matrix Spike

      A control matrix spike is prepared by spiking an aliquot of corn oil
      with the Aroclor 1254 spiking solution (25 ug/mL).   The Contractor
      shall perform one control matrix spike sample analysis for each Case
      received, or for each 20 single phase units,  or each 14 calendar day
      period during which single phase units in a Case were received (said
      period beginning with the receipt of the first sample in that Sample
      Delivery Group) whichever is most frequent.

3.     INITIAL CALIBRATION

3.1   Prior to analysis of samples,  it is necessary to run a successful
      calibration sequence using this sequence of injections:

                             TABLE E.I.  INITIAL CALIBRATION
                                                        No.  of Potential
                                              Aroclor  Quantitation Peaks
Injection No.   Concentration
                    1
                    2
                    3
                    4
                    5
                    6
                    7
                    8
                    9
                   10
                   11
                   12
                   13
                   14
                   15
                   16
                   17
                   18
                   19
                   20
                   21
                   22
                   23
                   24
                   25
                   26
     This  sequence will  establish potential quantitation peaks and  their
     retention times.  It will also establish a linear response calibration
     for each  analyte peak.
Blank
Resolution
Low

Check —
1221
Medium 1221
High
Low
Medium
High
Low
Medium
High
Low
Medium
High
Low
Medium
High
Low
Medium
High
Low
Medium
High
Low
Low
Blank
1221
1232
1232
1232
1242
1242
1242
1248
1248
1248
1254
1254
1254
1016/1260
1016/1260
1016/1260
Toxaphene
Toxaphene
Toxaphene
Pesticide A
Pesticide B
_

_
4
4
4
4
4
4
5
5
5
5
5
5
5
5
5
5/5
5/5
5/5
4
4
4
_
_
	
                                ARO E-47
                                                        Rev.  9/88

-------
3.2   Each Aroclor and Toxaphene calibration standard must be run at three
      concentrations.  The analytes and the concentrations required for the
      low point are given in Table E.2.  The midpoint concentrations are 10
      times the low point concentration, and the high point concentration is
      selected by the laboratory in the concentration range between 30 to 100
      times the low point concentration.  It is recommended that a
      concentration 100 times the CRQL be used for the high point
      calibration, provided that it lies within the linear range of the
      detector.  The high point calibration point defines the upper end of
      the concentration range for which the calibration is valid.
                   TABLE E.2.  LOW POINT AROCLOR CALIBRATION SOLUTIONS

                       Compound                CAS  Number  . Low Concentration

                  1.  Aroclor 1016              12674-11-2       50 ng/mL
                     Aroclor 1260              11096-82-5       50 ng/mL
                  2.  Aroclor 1221              11104-28-2       50 ng/mL
                  3.  Aroclor 1232              11141-16-5       50 ng/mL
                  4.  Aroclor 1242              53469-21-9       50 ng/mL
                  5.  Aroclor 1248              12672-29-6       50 ng/mL
                  6.  Aroclor 1254              11097-69-1       50 ng/mL
                  7.  Toxaphene                 8001-35-2      500 ng/mL
                  8.  Tetrachloro-meta-xylene    877-09-8       20 ng/mL
                  9.  Decaclorobiphenyl         2051-24-3       20 ng/mL

3.3   An on-scale chromatogram must be presented for each calibration run.

3.4   During the initial calibration sequence absolute retention times (RT)
      are determined for four or five major peaks of each Aroclor and
      Toxaphene.

3.5   If the mean deviation (M.D.) of the absolute retention time for the
      three calibration measurements for any peak exceeds 0.5 percent of the
      mean absolute retention time (RT) for that peak, the analytical system
     .is out of control and corrective action must be taken before collecting
      any data.
            M-D- "
                    13
                                 _
                    3  £  IRT, - RTI
                                                                        (3.1)
           — Absolute retention time of quantitation peak.
      RT   «- Mean absolute retention time of the quantitation
             peak.

3.6   A retention time window of +1.0 percent of the RT established during
      the initial calibration is calculated for each quantitation peak.

3.7   Three-point instrument calibration is required for each potential
      quantitation peak.  The laboratory has three choices on how to
      establish three-point calibration (sections 3.7.1 - 3.7.3).  Only one
                                 ARO E-48
Rev. 9/88

-------
      of the three calibration methods may be used to quantitate samples in
      any single run sequence.  Therefore, once a system is calibrated,
      laboratories may not change calibration techniques for samples until a
      new initial calibration is run.

      3.7.1   The laboratory may use  a mean calibration factor  (CF)
              determined  from the three concentrations but only if the % RSD
              for the three points is <15 percent.
                              n
                             i-1
       Response of
Peak Area (or Height)
                                                          of the Standard
                                             Mass Inj ected (ng)
                                    (3.2)
                            SD
                     %RSD - =  x 100
                            CF
                                 rn
                     Where SD - / Y  (CF, - CF)
                              A /l-l    *
                                       n-1
                                                        and n«=3
                                                                        (3.3)
                                    (3.4)
      3.7.2   The laboratory may use a calibration line drawn through all
     ,         three calibration points if the value for r  (the correlation
              coefficient from the linear regression calculation) is >0.975
              and if the\zero concentration intercept is <0.20 times the low
              point response for each analyte quantitated.

      3.7.3   Laboratories with electronic integrators or  data systems that
              automatically calculate calibration curves as line segments
        (      between calibration points may use two line  segment calibration
              curves for each quantitation peak of the Aroclor or Toxaphene.
              This technique may be used only if r (the correlation
              coefficient from the linear regression calculation) is >0.975
              for all three points for each peak and if the zero
              concentration intercept is <0.20 times the low point response
              for each Aroclor or Toxaphene peak quantitated.

3.8   Because they are potential method interferences,  mixtures of single
      component pesticides are injected as part of the calibration sequence
      to establish the RT  of individual pesticides.   Calibration factors are
      not calculated for the individual pesticide standards.
                                 ARO E-49
                                  Rev. 4/89

-------
3.9   TABLE E.3.  PESTICIDE MIXTURES
Individual
Standard
Mix A
Concentration
(ng/mL)
Individual
Standard
Mix fi
Concentration
(ng/mL)
      alpha-BHC                25.0
      Heptachlor               25.0
      gamma-BHC                25.0
      Endosulfan I             25.0
      Dieldrin                 50.0
      Endrin                   50.0
      p.p'-DDD                 50.0
      p,p'-DDT                 50.0
      Methoxychlor             50.0
      Tetrachloro-meta-xylene  20.0
      Decachlorobiphenyl       20.0
beta-BHC                25.0
delta-BHC               25.0
Aldrin                  25.0
Heptachlor epoxide      25.0
alpha-Chlordane         25.0
gamma-Chlordane         25.0
p,p'-DDE                50.0
Endosulfan sulfate      50.0
Endrin aldehyde         50.0
Tetrachloro-meta-xylene 20.0
Decachlorobiphenyl      20.0
Endrin Ketone           50.0
Endosulfan II           50.0
3.10  Sample analysis may not proceed until a satisfactory calibration has
      been demonstrated.

3.11  The more highly chlorinated Aroclor and Toxaphene components are more
      stable in the environment.  Therefore, the analyst should emphasize the
      later eluting peaks of a pattern in identifying and quantitating
      weathered Aroclors and Toxaphene

3.12  If more than one Aroclor is observed in a sample, or if an Aroclor and
      Toxaphene occur in the same sample, the laboratory must choose separate
      peaks to quantitate the different analytes.  A peak common to both
      analytes present in the sample must not be used to quantitate both
      Aroclors.

3.13  The surrogates must be added to all calibration standards.  The
      retention time shifts of the surrogates may not excede 0.5% for the
      analysis of any calibration standard.

3.14  Documentation for Initial Calibration

      3.14.1  Documentation of  the RT and  calibration  factors for potential
              quantitation  peaks on both columns must  be provided in tabular
              form.   Form VI will be used  to  report  the RT of each  potential
              quantitation  peak at each  of the  three concentrations.  The
              average RT and the relative  mean  deviation of  the  three
              measurements  shall also be reported.   Form VI  will be used to
              report  the CF for each peak  at  each concentration  injected, the
              concentrations injected, the average CF  and the relative
              standard  deviation of the  calibration  factors  at each of the
              three concentrations.
                                 ARO E-50
                              Rev.  9/88

-------
      3.14.2   Documentation of sequence of calibration standards  and their
               retention time shifts must be provided in tabular form on Form
               VIII.

      3.14.3   Documentation of the RT for pesticides on both columns must be
               provided in tabular form using Form IX.

3.15  Copies of Forms VI, VIII, and IX, as well as instructions for
      completion of those forms, are presented in Exhibit B of this  document.
      All chromatograms, as well as integration reports or data system
      printouts for calibration analyses must be submitted in hard copy with
      the data package.

4..   CONTINUED GC PERFORMANCE EVALUATION

4.1   Summary

      4.1.1    The  performance  of  the  GC must be monitored every 12 hours  by
               running  an instrument blank (Section 2.1)  and  an  evaluation
               standard.   The evaluation standard  is  used to  verify that  the
               RT's and CF's  of the analyte quantitation peaks have not
               changed  since  the initial calibration.

      4.1.2    Analysts are  cautioned  that analyzing  an instrument blank and
               an evaluation  mixture once every 12 hours is the  minimum
               contract requirement.   Highly complex  samples  or  unstable GC
               equipment may  cause  peaks from one  injection to be carried  over
               to the next.   It may be necessary to analyze these more often
               to avoid discarding  data.

      4.1,3    The  requirement  for  running the 12 hour  instrument blanks and
               evaluation mixture  is waived when no samples or spike  control
               matrix analyses  are  run during that 12-hour period.  After  a
               break in sample  data analysis,  a laboratory can resume the
               analysis of samples  and control matrix spikes  using the current
               initial  calibration  only after an acceptable evaluation mixture
               is analyzed (3.2).   If  a successful evaluation mixture cannot
               be analyzed after an interruption, an  acceptable  initial
               calibration must be  run before sample  data can be collected.

               NOTE: This section does not affect the requirement that all
               acceptable sample analyses must be bracketed by acceptable
               evaluation mixtures.

4.2   Performance Evaluation Standard
                                               •

      4.2.1    A performance evaluation standard must be  run  at  least once  in
               every 12 hours.  If  that analysis does not meet the criteria
               below, a second  injection of the same  standard at the  same
               concentration must be made  immediately.

      4.2.2    The  performance  evaluation standards injected at 12 hour (or
               less) intervals will be  rotated so as  to  include the Toxaphene
               and  common Aroclors at  low, medium,  and high concentrations
              according  to the following schedule:

                                ARO E-51                             Rev. 9/88

-------
                        TABLE  E.4.  12 HOUR EVALUATIONS
        Concentration

            low
            low
            low
            low
            medium
            medium
            medium
            medium
            high
            high
            high
            high
            Repeat
                                 Compound

                                Toxaphene
                                   1248
                                   1254
                                1016/1260
                                Toxaphene
                                   1248
                                   1254
                                1016/1260
                                Toxaphene
                                   1248
                                   1254
                                1016/1260
Hour

 12
 24
 36
 48
 60
 72
 84
 96
 108
 120
 132
 144
4.2.3   For each of  the  four or  five potential quantitation peaks,  the
        RT in  the standard must  be within the retention  time window of
        ±1.0 percent of  the mean RT calculated during  the  initial
        calibration. .

4.2.4   For each of  the  four or  five potential quantitation peaks,  the
        response (area or height) must be within  20.0  percent  (±20.0
        RPD) of the  mean response obtained  during the  initial
        calibration.
      RPD  -
                     CF   CF
                                 100
                                                                  (4.1)
                        CF
4.2.5
CF - Average calibration factor from initial
     calibration (Equation 3.2).

CFg — Calibration factor from current
      evaluation mixture.

The retention time shifts of the surrogates in the evaluation
standard must not excede 0.5%.
4.2.6    If the performance evaluation mixture  does  not meet all  of the
         criteria listed above,  the GC system is  out of control and
         appropriate corrective  action must then  be  taken before
         additional data are collected.   After  corrections have been
         made,  the initial calibration must be  run successfully.   All
         samples analyzed since  the previous acceptable evaluation must
         be reinjected at no additional cost to the  Agency.  The
         laboratory manager, or  his designate,  must  address problems and
         solutions in the narrative.
                           ARO E-52
                                                        Rev. 9/88

-------
4.3   Documentation

      4.3.1   Documentation of an acceptable performance evaluation mixture
              analysis  on  each column shall be submitted on Form VII for each
              12 hours  of  analysis.  In addition, each analysis of a
              performance  evaluation mixture must be listed on the analytical
              sequence  given on Form VIII regardless of whether or not it was
              acceptable.

      4.3.2   A hard copy  of all chromatograms of performance evaluation
              analyses, as well as integration reports or data system
              printouts, must be submitted with the data package.

4.4   Instructions for completing the data reporting forms are contained in
      Exhibit B.

5.    BLANK ANALYSIS

5.1   Two types of blanks are required as  part of this  protocol.   Method
      blanks, which provide a measure of total laboratory contamination,  and
      instrument blanks, which provide a measure  of instrument contamination
      and sample carry over between injections.

5.2   Method Blank

      5.2.1   A method blank of corn oil (500 mg) must be extracted and
              cleaned up using the protocols given in Exhibit D.

      5.2.2   A method blank analysis must be performed for each twenty
              single phase units in each sample delivery group.  It is
              suggested that blanks be run more often whenever particularly
              dirty samples are analyzed.

      5.2.3   If only a partial.set of samples require sulfur cleanup,  then
              two method blanks are required, [one that is shaken with
              mercury (or  copper) and one that is not].

      5.2.4   The method blanks must be injected directly after a sample in
              the run sequence.  They may not be run immediately after an
              instrument blank.

      5.2.5   The method blank may not contain more than the contract
              required quantitation limit of any Aroclor or Toxaphene
              quantitation peak.  If a method blank exceeds the criterion,
              the analytical system is out of control.   The source of the
              contamination must be investigated and appropriate corrective
              measures must be taken.  All samples processed with a method
              blank that is out of control (i.e., contaminated) must be re-
              extracted, cleaned up again, and reanalyzed at no additional
              cost to the  Agency.  The laboratory manager, or his designate,
              must address problems and solutions in the narrative (Exhibit
              B>-

      5.2.6   The retention time shifts of the surrogates in the method blank
              must not excede 0.5%.

                                ARO  E-53                              Rev.  9/88

-------
5.3   Instrument Blank
      5.3.1   An instrument blank is a hexane solution containing 20 ng/mL of
              both of the surrogates.  An acceptable instrument blank must be
              run at least once every 12 hours.  An instrument blank must be
              run immediately prior to the performance evaluation mixture in
              the run sequence (see Exhibit D) .  To avoid the necessity of
              reanalysis, it is recommended that additional blanks be
              analyzed whenever particularly dirty samples are analyzed.

      5.3.2   An acceptable instrument blank analysis must demonstrate that
              no potential quantitation peak of an Aroclor or Toxaphene is
              detected at greater than 0.5 times the CRQL.  If quantitation
              peaks identified during the initial calibration are detected at
              greater than half the CRQL, all data collection must be stopped
              and corrective action taken.  Data for samples with analytes
              detected at >CRQL which were analyzed between the last
              acceptable instrument blank and the unacceptable blank must be
              considered suspect.  An acceptable instrument blank must be run
              before additional data is collected.  After an acceptable
              instrument blank is run, all samples with Aroclors or Toxaphene
              detected at levels >CRQL analyzed after the last acceptable
              instrument blank must be reinjected during a valid run sequence
              and reported at no expense to the Agency.

      5.3.3   If an instrument blank exceeds the 0.5 x CRQL criteria, the
              Contractor must consider the analytical system out of control.
              The source of the contamination must be investigated and
              appropriate corrective measures must be taken.  The laboratory
              manager, or his designate, must address problems and solutions
              in the narrative (Exhibit B) .

      5.3.4   The retention time shifts of the surrogates in the instrument
              blank must not excede 0.5%. •

5 . 4   Documentation

      5.4.1   Results of method blank analysis must be reported using Form I
              (Aroclor/Toxaphene Analysis Data Sheet) .  In addition, the
              samples associated with each method blank must be summarized on
              Form IV.  Specific instructions for the completion of these
              forms can be found in Exhibit B.

      5.4.2   Results of instrument blank analyses must be reported using
              Form I.  In addition, the samples associated with each
              instrument blank must be summarized on Form IV.  Specific
              instructions for the completion of these forms can be found in
              Exhibit B.
      5.4.3   In addition, each analysis of a method blank or instrument
              blank must be reported on Form VIII for each analytical
              sequence .
                                 ARO E-54
Rev. 9/88

-------
      5.4.4    Chromatograms, as well as  integration reports or data system
               printouts  for calibration  analyses must be submitted in hard
               copy with  the data package.

      5.4.5    The Contractor will report ALL sample concentration data
               UNCORRECTED  for blanks.  It shall be the responsibility of the
               EPA evaluator, or data auditor, to correct analyte
               concentrations for concentrations detected in the blank(s).  It
               is the  Contractor's responsibility to ensure that the proper
               number  of  blanks are analyzed and,that the data are properly
               reported.

6.    CONTROL MATRIX SPIKE ANALYSIS

6.1   Control Matrix Spike Requirements

      6.1.1    In order to  evaluate the effect of the sample matrix upon the
               analytical methodology, the EPA has specified that a solution
               containing Aroclor 1254 at 25 ug/mL) be used to spike a control
               matrix  of  corn oil (500 mg).

      6.1.2    A control  matrix spike must be analyzed once for each Case, or
               for every  20 single phase  units.

6.2   Calculations

      6.2.1    The recovery of each Aroclor 1254 in the control matrix spike
               is calculated using Equation 6.1.

               Control
               Matrix
               Spike      -         SSR    x 100   ,fi ,»
               Percent             SA                '
               Recovery

               Where:

               SSR  - Spike Sample Results
               SA   - Spike Added from Spiking Mix

6.3   Documentation

      6.3.1    The concentration of nonspike target compounds in the control
               matrix spike must be reported on Form I.   The quantitation of
               each Aroclor 1254 peak on  both columns must be reported on Form
               X.  The recovery of Aroclor 1254 must be reported on Form III.
               Specific instructions on the completion of these forms may be
               found in Exhibit B.

      6.3.2    Chromatograms as well as integration reports or data system
               printouts  of control matrix spike analyses, will be submitted
               in hard copy with the data package.
                                 ARO E-55                              Rev. 4/89

-------
7.
SAMPLE CLEANUP
7.1.  Summary

      7.1.1   The cleanup of sample requires Diol cartridges for all samples.

      7.1.2   Every lot number of Diol cartridges must be tested by the
              following procedure before they are used for sample cleanup.
              Add 0.5 mL of Aroclor, midpoint concentration (described in
              Exhibit D, Section 5.5.1) to 4 mL of hexane, then reduce the
              final volume to 0.5 mL using nitrogen  (Exhibit D, Section 5.4).
              Place the mixture onto the top of a prewashed Diol cartridge,
              and elute it with 9 mL of hexane/acetone  [(90:10)(V/V)].
              Adjust the final volume to 10.0 mL and analyze by GC/ECD.

      7.1.3   The recovery of Aroclor 1254 must be determined  for evaluation
              and reporting purposes.  The lot of Diol cartridges is
              acceptable if the Aroclor is recovered at  80 to  110 percent.

7.4   Documentation

      7.4.1   Documentation of the Diol cartridge performance  will be
              provided on Form XI by reporting the recovery of Aroclor 1254.

8.    SAMPLE ANALYSIS SUMMARY

8.1   General

      8.1.1   This  section DOES NOT  replace  or supersede the specific
              analytical methods or QA/QC  activities described in previous
              sections.  The  intent  of this  subsection  is to provide the
              contractor laboratory with a BRIEF summary of QC activities
              involved with sample analysis  to help  the  contractor laboratory
              meet  specific reporting and  deliverables  required by this
              contract.

      8.1.2   Samples may only be analyzed upon successful completion of the
              initial QC activities.  The  laboratory must run  an evaluation
              mixture and an  instrument blank  (described in Section 2.1) at
              least every 12 hours.  The laboratory  may  continue to
              quantitate data using  the initial calibration until an
              instrument blank shows greater than half  the contract minimum
              level of any Aroclor or Toxaphene, an  unacceptable evaluation -
              mixture is analyzed, or the  calibration factor of a
              multicomponent analyte changes by more than 20 percent.  Any
              major system maintenance, such as installation of a new column
              or, changing or cleaning the detector  will also  necessitate
              reanalysis of the initial calibration  sequence.

      8.1.3   All acceptable data must be bracketed  with an acceptable
              instrument blank and an acceptable evaluation mixture.  Any
              samples analyzed which do not  meet this criteria must be
              reinjected and reported at no  expense  to  the Agency.
                                 ARO E-56
                                                                 Rev.  9/88

-------
      8.1.4   At  least  one  acceptable evaluation mixture must be run every  12
              hours.  If  an unacceptable evaluation run is made, the
              laboratory  must  run a  second evaluation mixture immediately.
              If  the  results of  two  successive evaluation runs do not meet
              the specifications of  Section 4.2, then all analyses since the
              last acceptable  evaluation mixture are not valid.  They must  be
              reinjected  after the GC is inspected for malfunction and
              corrections made as appropriate.   Loss of sensitivity or
              resolution  could result from a damaged septum or column.  High
              detector  standing  current (background) could result from
              contaminated  carrier gas or from a dirty detector.

      8.1.5   An  acceptable instrument blank analysis must demonstrate that
              no  potential  quantitation peak of an Aroclor or Toxaphene is
              detected  at greater than 0.5 times the CRQL.  If quantitation
              peaks identified during the initial calibration are detected  at
              greater than  half  the  CRQL, all data collection must be stopped
              and corrective action  taken.  Data for samples with analytes
              detected  at >CRQL which were analyzed between the last
              acceptable  instrument  blank and the unacceptable blank must be
              considered  suspect.  An acceptable instrument blank must be run
              before  additional  data is collected.  After an acceptable
              instrument  blank is run, all samples with Aroclors or Toxaphene
              detected  at levels >CRQL analyzed after the last acceptable
              instrument  blank must  be reinjected during a valid run sequence
              and reported  at  no expense to the Agency.

      8.1.6   At  least  one  method blank must be analyzed for each 20 single
              phase units analyzed.  If any Aroclor or Toxaphene is detected,
              in  a method blank,- at  more than the CRQL, all sample data
              collection  must  stop.  The laboratory must then immediately run
              an  instrument blank to demonstrate that the instrument is not
              contaminated.  When an acceptable instrument blank has been
              analyzed, the same method blank must be reinjected.  If the
              method  blank  is  unacceptable after the second injection, all
              single  phase  units that were analyzed since the last acceptable
              method  blank  must be re-extracted, cleaned up again and
              analyzed  at no expense to the Agency.

8.2   Sample Dilution

      8.2.1   The protocol  is  intended to achieve the quantitation limits
              shown in  Exhibit C whenever possible.  Whenever sample
              chromatograms have interfering peaks, a high baseline, or off-
              scale peaks,  the samples must be reanalyzed following further
              dilution.   Samples which cannot be made to meet the
              specifications given in this section after dilution and cleanup
              (sulfuric acid,  permanganate, Diol cartridge and mercury
              removal)  are  discussed in detail in the Case Narrative and do
              not  require further analysis.  No limit is placed on the number
              of  re-extractions and  cleanups of samples that may be required
              because of contaminated method blanks.
                                 ARO E-57
Rev. 9/88

-------
8.2.2
8.2.3
              The sample must be analyzed at the most concentrated level
              consistent with achieving satisfactory chromatography (defined
              below).  If dilution is employed solely to bring a peak within
              the calibration range or produce an on-scale chromatogram for
              the multicomponent analyte, the results for both a more and a
              less concentrated dilution must be reported.  The resulting
              changes in quantitation limits and surrogate recoveries must
              also be reported for the dilute samples.

              If the laboratory has reason to believe that diluting the final
              volume will be necessary, an undiluted run may not be required.
              If an acceptable chromatogram (as defined in Section 5.) is
              achieved with the diluted final volume, an additional analysis
              at 10 times the concentration of the diluted material must be
              injected and reported with the sample data.

              The response of peaks in the sample must be >25 percent of full
              scale to allow visual pattern recognition of multicomponent
              analytes.

              An on-scale chromatogram(s) of all identified peaks must be
              presented with the sample data.

              Chromatographic data may be replotted electronically in order
              to get an on-scale chromatogram, except when the off-scale
              sample peaks are larger than the high point calibration peaks.
              In that case, the samples must be diluted and reinjected.

              The peak response of sample peaks on the replotted chromatogram
     	__^    must'be >25 percent of full scale to allow visual pattern
       ~recognition of multicomponent analytes.

8.3   Data Acceptance Criteria

      8.3.1   Reportable data for a sample must include a. chromatogram with a
              baseline.which returns to below 50 percent of full scale before
              the elution time of Aroclor 1221 and to below 25 percent of
              full scale after Aroclor 1221 and before the elution time of
              Decachlorobiphenyl.

      8.3.2   If dilution has been applied and no peaks are detected above 25
              percent of full scale, analysis of a more concentrated extract
              is required.

      8.3.3   Reportable sample data must include chromatogram(s) with all
              Aroclor and Toxaphene quantitation peaks in the linear range of
              the quantitation curve determined by the initial calibration.

      8.3.4   Reportable sample data must include chromatogram(s) with all
              Aroclor and Toxaphene peaks on-scale.

              NOTE: If more than one chromatogram is required to satisfy the
              criteria for a sample, the results of all chromatograms must be
              reported.  These requirements apply to both columns.
8.2.4
8.2.5
8.2.6
8.2.7
                           ARO E-58
                                                                      Rev. 9/88

-------
      8.3.5   The retention time shifts for the surrogates in any sample must
              not excede 0.5%.  If the surrogates have been diluted out, so
              that no retention time shift can be calculated, report the
              percent difference as "dil" (see Forms Instructions), and
              explain in the Case Narrative.

8.4   Quantitation of Analyses

      8.4.1   Analytes can be quantitated using either a modern electronic
              integrator or with a laboratory data system.  The analyst can
              use either peak height or peak area as the basis for
              quantitation.  The use of an electronic integrator or a
              laboratory data system is required.

      8.4.2   The chromatograms of all samples must be reviewed by a
              qualified pesticide/PCB analyst before they are reported.

      8.4.3   Using an electronic integrator, one of three calibration
              techniques may be employed (Section 3.7).  It is the
              responsibility of the analyst to set the integration parameters
              such that off -scale chromatograms are within the dynamic range
              of the instrument.  The analyst should also check for data
              flags generated by the instrument that indicate improper
              quantitation of peaks prior to reporting data to the EPA.

      8.4.4   In order to be quantitated,  the dete'ctor response (peak area or
              peak height) of all analytes must lie between the responses of
              the low and high concentrations in the initial calibration.  If
              the analytes are detected below the CRQL, they are reported
              with the appropriate flags (See Exhibit B) .   If they are
              detected at a level greater than the high calibration point,
              the sample must be diluted either to a maximum of 1:100,000, or
              until the response is within the linear range established
              during calibration.

      8.4.5   The concentration of the Aroclor and Toxaphene quantitation
              peaks are calculated using the following equations:
                     Concentration ug/L
                                             (AX)
                                           (CT)(Vi)(Wx)                  (8.1)

              Where:

                     A  - Response for the peak to be measured.
                     5? - Calibration factor from the initial calibration.
                     Vt - Volume of total waste dilution (uL) (take into
                          account any dilution).
                     Vj  =• Volume of waste dilution injected (uL) .
                     Wx - Weight of waste diluted (gm).
                                ARO  E-59                              Rev. 9/88

-------
      8.4.6   The the recoveries of the surrogates are calculated using the
              following equations:
                     Percent Recovery
SA - SR
   SA
x 100%
                                                                        (8.2)
                     SA - Concentration of surrogate added
                     SR — Concentration of surrogate recovered

      8.4.7   The laboratory will quantitate each of the selected Aroclor or
              Toxaphene peaks individually, and determine an average
              concentration from all of the selected peaks.  Quantitation is
              performed on both columns, and the lower value is reported on
              Form I.  See Exhibit B for instructions on completing Form I
              and Form X.

      8.4.8   Detected Aroclors and Toxaphene must be reported as: (1) a
              concentration between the CRQL and 10  x CRQL, or (2) as an
              estimated value below the CRQL.  (See Exhibit B.)

      8.4.9   The choice of the peaks used for Aroclor or Toxaphene
              quantitation and recognition of those peaks may be complicated
              by the environmental alteration of the Aroclor or Toxaphene,
              and by the presence of coeluting analytes or matrix
              interferences.

      8.4.10  The more highly chlorinated components of the Aroclors and
              Toxaphene are more stable in the environment.  Therefore, the
              analyst should emphasize the later eluting peaks of a pattern
              when identifying and quantitating weathered Aroclors or
              Toxaphene.

      8.4.11  If more than one Aroclor is observed in a sample or if an
              Aroclor and Toxaphene are both observed in a sample the
              laboratory must choose separate quantitation peaks for the
              different Aroclors and Toxaphene.  A peak common to both
              analytes in the sample must not be used to quantitate both
              compounds.

8.5   Documentation

      8.5.1   Refer to Exhibit B of this document for specific details on
              contract deliverable and reporting formats.  Exhibit B contains
              specific instructions for completing all required forms, as
              well as detailed itemization of deliverables and reporting
              requirements.

9.     Surrogates

9.1   Requirements

      9.1.1   Tetrachloro-meta-xylene and Decachlorobiphenyl must be added to
              each sample, control matrix spike, and blank analyzed as part
              of this protocol.
                                 ARO E-60
                          Rev.  4/89

-------
      9.1.2   The recoveries of the surrogates are calculated using equation
              8.2.

9.2   Documentation

      9.2.1   The recoveries of the surrogates are reported for each sample,
              diluted sample, and control matrix spike on Form II.

      9.2.2   Complete  instructions for the completion of Form II are given
              in Exhibit B.

10.   GC Maintenance

10.1  Laboratory Responsibility

      10.1.1  It is critical that the gas chromatographs used for these
              analyses  be maintained properly and that all manufacturers'
              recommendation be followed.  Although not addressed in this
              document, it  is expected that each analytical laboratory will
              maintain  performance checks on their instruments, including
              monitoring detector standing current and monitoring the number
              of theoretical plates per column.

10.2  Suggested Maintenance

      10.2.1  GC Columns -  When degradation in column performance is
              observed, it  is usually the result of a build up of
              nonvolatiles  on the head of the column.  These can best be
              removed by cutting off 1-2 feet of the injector end of the
              column.   The  analyst must cool the column(s) to room
              temperature before exposing them to air or the column(s) will
              oxidize and lose some ability to resolve analytes.

      10.2.2  Septa - must  be replaced regularly to prevent degradation of
              chromatographic performance and destruction of the columns.
              They must be  replaced at least once per day whenever data are
              collected.

      10.2.3  Carrier Gas - should be high purity, oxygen, and water free.
              The use of in-line water and oxygen traps is recommended even
              if ultra-high purity gas is used.

      10.3.4  Electron  Capture Detector - The condition of the detector can
              be monitored  by its standing current.  A high standing current
              usually indicates contaminated carrier gas, a high column bleed
              rate, or  a dirty detector.  Changing the gas filters, oxygen
              traps, or the column may correct the problem.  If it does not,
              raise the detector temperature overnight to bake out the
              detector.

      10.2.5  NOTE: Do  not  collect data while cleaning the detector.
                                 ARO E-61
Rev. 9/88

-------
      10.2.6  More serious contamination will require steam cleaning the
              detector or detector repair by the manufacturer.  Steam
              cleaning or detector repair will necessitate recalibration of
              the instrument.

10.3  Documentation

      10.3.1  All major instrument maintenance or repair will be recorded,
              dated, and initialed in an instrument log that will be made
              available to auditors during on-site inspections.

11.    Solvents and Reagents

11.1  Aliquots of 100 mL of all lots of solvents used for the analysis of
      samples by this method will be concentrated,  exchanged to a final
      volume of 1.0 mL of hexane and analyzed by GC/ECD.   The solvent lot is
      acceptable only if no analyte is detected at >0.5 CRQL.

11.2  Commercial standards must be diluted and analyzed by GC/ECD to
      demonstrate that they give acceptable retention times and that they
      give a response between 95 to 105 percent of EPA repository standards.

11.3  Documentation

      11.3.1  Chromatograms  for  all lots of solvents, and commercial
              standards must be  on file at the laboratory.  They are not a
              contract deliverable.
                                 ARO E-62                              Rev.  9/88

-------
                                     SECTION  IV
                                ANALYTICAL STANDARDS
A.   The Environmental Protection Agency's Quality Assurance Materials Bank will
     supply primary standards (calibration standards,  surrogate standards,  control
     matrix spiking standards, and internal standards),  contingent upon their
     availability,  for traceability and quantitative verification of Contractor
     standards.   It is emphasized that these primary standards are for traceability
     only.  There are insufficient quantities to have  these available to serve as
     working standards.  The Contractor laboratory is  responsible for preparing its
     own working standards from commercial sources.

B.   Caution should be exercised in the mixing of these  standards, particularly the
     multicomponent standards.  Chemical reactions such  as acid/base reactions,
     Schiff base formations (reactions of aldehydes  and  ketones with primary
     amines), hydrolysis, isotopic exchange, and others  may occur.

C.   EPA contract laboratories can call or write directly to the QAMB (address and
     phone number on the following request form) to  obtain reference standards.
     Standards will be provided based on the reasonableness of the request  and
     their availability.  Any request from a commercial  laboratory that is  not
     currently under contract to EPA will be denied.

D.   Upon award of a contract, a list of available standards will be provided by
     EMSL-LV upon request.
                                    E-63
Rev. 9/88

-------
             QUALITY ASSURANCE MATERIALS BANK
        RE0U EST FOR REFERENCE STANDARDS
  MAtL TO:  U.S. Environmental Protection Agency
          Quality Assurance Materials Bank (MD-8)
          Research Triangle Park. NC* 27711 USA
Telephone: Requests ONLY: (919) 541-4019
CFTS) 629-4019
• ••• ' > Information: (702)545-2690
OTS) 545-2690
• •:•:••'.:•.- :•.'•'; Technical Assistance: (919)541-3951
; |RS) 629-3951

Number
Acquired




*





••" •"--•. •




Codi
Number


"



•:;-.;.-.;. •• - -,-.'. -
. . '

•
? '•:?.''-.? ' •
. .' •

-

OncltcaiMitlUccnMd
Deleft of XH£BRM
m

v«
rificd
IMS ILOOC TOM AGENCY USC ONLY

Campowddl

*

„


•••••;....-,-.. -,--.-: . .. - • ••• .. '


•

.' •' :-:.'•-:••- ' -



*
So,v,«













" . -

Purity






- ,v,r<^..








• Conctmratton















-~mi ro~o ni -nifir- 'nfifti—IT—F QiHTrnrt n 11 inma unrun- ""nin-i'i tMtMaofwMf W
                   Nun* *»d Addrvn of Ub
-------
                                     SECTION V
                          LABORATORY EVALUATION PROCEDURES


This section outlines the procedures which will be used by the Project Officer or
his/her authorized representative during the contract period of performance to
conduct laboratory audits to determine the Contractor's continuing ability to meet
the terras and conditions of this contract.  The evaluation process incorporates
two major steps:  1) evaluation of laboratory performance, and 2) on-site
inspection of the laboratory to verify continuity of personnel, instrumentation
and quality control requirements of the contract.  The following is a description
of these two steps.

Part 1 - Evaluation of Laboratory Performance

1.    Performance Evaluation Sample Analysis

1.1   A Performance Evaluation (PE) sample set may be sent to a participating
      laboratory on a quarterly basis to verify the laboratory's continuing
      ability to produce acceptable analytical results.  These samples will be
      provided either as single blind (recognizable as a PE material and of
      unknown composition) or double blind (not recognizable as a PE material and
      of unknown composition).

      If received as a single blind, the contractor is required to submit PE
      sample data in a separate SDG package in accordance with Delivery Schedule
      requirements for sample data.  PE samples received as double blind would be
      treated as routine samples and data would be submitted in the SDG
      deliverables package per normal procedure.

1.2   When the PE data are received, results will be scored routinely for
      identification and quantitation.  Results of these scorings will be provided
      to the Contractor.  The government may adjust the scores on any given PE
      sample to compensate for unanticipated difficulties with a particular
      sample.

1.3   If a laboratory performs  unacceptably, the laboratory will be immediately
      notified by the Project Officer.  A laboratory so notified may expect, but
      the government is not limited to, the following actions:  a site visit, a
      full data audit,  and/or laboratory analysis of a second PE sample.  Failure
      by the laboratory to take corrective actions and/or failure of two
      successive PE sample analyses will require that the laboratory discontinue
      analysis of samples until such time as the Project Officer has determined
      that the laboratory may resume analyses.

2.    Organic Data Audit

2.1   Organic data audits are conducted on CLP Contractor's Reporting and
      Deliverables packages by  EMSL/LV.  The organic data audit provides the
      Agency with an in-depth,inspection and evaluation of the Case data packages
      with regard to achieving  QA/QC acceptability.
            f

                                     E-65                            Rev. 9/88

-------
Part 2 - On- Site Laboratory Evaluation

2.    The on-site laboratory evaluation helps to ensure that all the necessary
      quality control is being applied by the Contractor in order to deliver a
      quality product.

2.1   Quality assurance evaluations allow the evaluators to determine that:

      2.1.1   The organization and personnel are qualified  to perform assigned
              tasks,

      2.1.2   Adequate  facilities and  equipment are available,

      2.1.3   Complete  documentation,  including chain-of-custody of  samples  is
              being  implemented,

      2.1.4   Proper analytical methodology  is being used,

      2.1.5   Adequate  analytical Quality Control, including reference  samples,
              control charts, and documented corrective action measures,  is  being
              provided,  and

      2.1.6   Acceptable data handling and documentation  techniques  are being
              used.

2.2   The on-site visit also serves as a mechanism for discussing weaknesses
      identified through the Performance Evaluation sample analysis or  through
      Contract Compliance Screening or other review of data deliverables.  Lastly,
      the on-site visit allows the evaluation team to determine if the  Contractor
      has implemented the recommended and/or required corrective actions, with
      respect to quality assurance, made during the previous on-site visit.
                                     E-66
Rev. 9/88

-------
             EXHIBIT F
CHAIN-OF-CUSTODY, DOCUMENT CONTROL,




 AND STANDARD OPERATING PROCEDURES
                                               Rev.  9/88

-------

-------
1.    Sample Chaln-o£-Custody

      A sample is physical evidence collected from a facility or from the
      environment.  An essential part of hazardous waste investigations is
      that samples and data may be used as evidence in EPA enforcement
      proceedings.  To satisfy enforcement uses of the data, the following
      chain-of-custody procedures have been established.

1.1   Sample Identification

      To assure traceability of samples while in possession of the
      Contractor, a method for sample identification shall be developed and
      documented in laboratory Standard Operating Procedures (SOPs)  (see
      Section 3).  Each sample or sample preparation container shall be
      labeled with a unique number identifier (or the SMO number).  • This
      identifier shall be cross-referenced to the sample tag number and the
      SMO number.  There shall be a written description of the method of
      assigning this identifier and attaching it to the sample container
      included in the laboratory SOPs.

      1.2.1   A  sample  is under custody  if:

              1.2.1.1    It is in your actual possession,

              1.2.1.2    It is in your view after being in your physical
                         possession,

              1.2.1.3    It was in your possession and then you locked or
                         sealed it up to prevent tampering, or

              1.2.1.4    It is in a'secure area.

      1.2.2   Upon  receipt of  the samples  in custody,  the Contractor shall
              inspect the shipping  container and sample bottles and shall
              document  receiving  information as specified in  Section 3.2.
              The sample custodian  or a  designated representative  shall
              sign  and  date all appropriate receiving  documents at the  time
              of receipt (i.e., EPA chain-of-custody forms, traffic
              reports,  airbills,  etc.).  The Contractor shall contact SMO
              if documents are absent, information on  receiving documents
              does  not  agree,  custody seals are not intact, or the sample
              is not in good condition.  The Contractor shall document
              resolution of any discrepancies, and this documentation shall
              become a  part of the  permanent case file.

      1.2.3   Once  samples have been accepted by the laboratory, checked,
              and logged in, they must be maintained in accordance with
              custody and security  requirements specified in  3.3.

2.    Documen^Control Procedures

      The goal of the laboratory document control program is to assure  that
      all documents  for  a specified case will  be accounted for when the
      project is  completed.   Accountable documents  used by contract

                                   F-2                              Rev. 9/88

-------
      laboratories shall include,  but not be limited to,  logbooks,  chain-
      of-custody records, sample work sheets,  bench sheets,  and other
      documents relating to the sample or sample analyses.   The following
      document control procedures have been established to assure that all
      laboratory records are assembled and stored for delivery to EFA or
      are available upon request from EPA prior to the delivery schedule.

2.1   Preprinted Data Sheets and Logbooks

      Preprinted data sheets shall contain the name of the laboratory and
      be dated and signed by the analyst or individual performing the work.
      All documents produced by the laboratory which are  directly related
      to the preparation and analysis of EFA samples shall become the
      property of the EPA and shall be placed in the case file.  For that
      reason, all observations and results recorded by the laboratory but
      not on preprinted data sheets are entered into permanent laboratory •
      logbooks.  The person responsible for the work shall sign and date
      each entry and/or page in the logbook.  When all data from a case is
      compiled, copies of all EPA case-related logbook entries shall be
      included in the documentation package.  Analysts'  logbook entries
      must be in chronological order and shall include only one case per
      page.  Instrument run logs shall be maintained so as to enable a
      reconstruction of the run sequences of individual instruments.

      Because the laboratory must provide copies of the instrument run logs
      to EPA, the laboratory may exercise the option of using only
      laboratory or SMO sample identification numbers in the logs for
      sample ID rather than government agency or commercial client names.

      Using laboratory or SMO sample IDs only in the run sequences will
      assist the laboratory in preserving the confidentiality of commercial
      clients.

2.2   Error Correction Procedure

      All documentation in logbooks and other documents shall be in ink.
      If an error is made, corrections shall be made by crossing a line
      through the error and entering the correct information.  Changes
      shall be dated and initialed.  No information shall be obliterated or
      rendered unreadable.

2.3   Consistency of Documentation

      Before releasing analytical results, the laboratory shall assemble
      and cross-check the information on sample tags, custody records, lab
      bench sheets, personal and instrument logs, and other relevant data
      to ensure that data pertaining to each particular sample or case is
      consistent throughout the case file.
                                   F-3
Rev. 9/88

-------
 2.4    Document  Numbering and  Inventory Procedure

       In  order  to provide document accountability of  the completed analysis
       records,  each  item in a case shall be  inventoried and assigned a
       serialized number  and identifier associating it to the case and
       Region.

       Case #  -  Region  -  Serialized number  (For example:  75-2-0240)

       The number of  pages of  each item must  be accounted for if each page
       is  not  individually numbered.  All documents relevant to each case,
       including logbook  pages, bench sheets, mass spectra, chromatograms,
       custody records, library search results, etc.,  shall be inventoried.
       The laboratory shall be responsible  for ensuring that all documents
       generated are  placed in the file for inventory  and are delivered to
       EPA.  Figure 1 is  an example of a document inventory.

 2.5    Shipping  Data  Packages  and Case Files

            The Contractor shall have written procedures to document
            shipment of  deliverables packages to the  recipients.  Case File
            Purge shipments require custody  seals on  the container(s)
            placed such  that  it cannot be  opened without damaging or
            breaking the seal.  The Contractor shall  also document what was
            sent, to whom, the date, and the method (carrier) used.
.3.    Standard Operating Procedures

      The Contractor must have written standard operating procedures  (SOPs)
      for (1) receipt of samples,  (2) maintenance of custody, (3) sample
      storage, (4) tracking the analysis of samples, and (5) assembly of
      completed data.

      An SOP is defined as a written narrative step-wise description of
      laboratory operating procedures including examples of laboratory
      documentation.  The SOPs must accurately describe the actual
      procedures used in the laboratory, and copies of the written SOPs
      shall be available to the appropriate laboratory personnel.  These
      procedures are necessary to ensure that analytical data produced
      under this contract are acceptable for use in EPA enforcement case
      preparation and litigation.  The Contractor's SOPs shall provide
      mechanisms and documentation to meet each of the following
      specifications and shall be used by EPA as the basis for laboratory
      evidence audits.

3.1   The Contractor shall have a designated sample custodian responsible
      for receipt of samples and have written SOPs describing his/her
      duties and responsibilities.

3.2   The Contractor shall have written SOPs for receiving and logging in
      of the samples.   The procedures shall include but not be limited to
      documenting the following information:
                                   F-4
Rev. 9/88

-------
            o  Presence or absence of EPA chain-of-custody forms

            o  Presence or absence of airbills

            o  Presence or absence of EPA Traffic Reports or  SAS packing
               lists

            o  Presence or absence of custody seals  on  shipping and/or
               sample  containers  and their condition

            o  Presence or absence of sample tags

            o  Sample  tag ID numbers if not recorded on the chain-of-
               custody record(s)  or packing list(s)

            o  Condition of the shipping container

            o  Condition of the sample bottles

            o  Verification of agreement or nonagreeraent of information  on
               receiving documents

            o  Resolution of problems or discrepancies  with the Sample
               Management Office

3.3   The Contractor shall have written SOFs  for maintenance of the
      security of samples  after log-in and shall demonstrate security of
      the sample storage  and laboratory areas.   The SOPs shall specifically
      include descriptions of all storage areas  for EPA samples in the
      laboratory,"and^"steps taken to prevent  sample contamination.   The
      SOPs shall include  a list of authorized personnel who have access or
      keys to secure storage areas.

3.4   The Contractor shall have written SOPs  for tracking the work
      performed on any particular sample.   The tracking SOP shall include
      the following:

      3.4.1   A description of the documentation used to record sample
              receipt, sample storage, sample transfers, sample
              preparations, and sample analyses.

      3.4.2   A description of the documentation used to record instrument
              calibration and other QA/QC activities.

     * 3.4.3   Examples of the document formats  and laboratory documentation
              used in  the sample  receipt, sample storage, sample transfer,
              and sample analyses.

3.5   The Contractor shall have written SOPs  for organization and assembly
      of all documents relating to each EPA case, including technical and
      managerial review.   Documents shall be filed on a Case-specific
      basis.  The procedures must ensure that all documents including
      logbook pages, sample tracking records,  chromatographic charts,
      computer printouts,  raw data summaries,  correspondence,  and any other
                                    F-5
Rev. 9/88

-------
      written documents having reference to the Case are compiled in one
      location for submission to EPA.  The system must include a document
      numbering and inventory procedure.

3.6   The Contractor shall have written SOPs for laboratory safety.

3.7   The Contractor shall have written SOPs for cleaning of glassware used
      in preparing and analyzing samples under this contract.

3.8   The Contractor shall have SOPs for traceability of standards used in
      sample analysis QA/QC.

4.    Handling of Confidential Information

      A Contractor conducting work under this contract may receive EPA-
      designated confidential information from the agency.  Confidential
      information must be handled separately from other documentation
      developed under this contract.  To accomplish this, the following
      procedures for the handling of confidential information have been
      established.

4.1   All confidential documents shall be under the supervision of a
      designated document control officer (DCO).

4.2   Confidential Information

      Any samples or information received with a request of.confidentiality
      shall be handled as "confidential."  A separate locked file shall be
      maintained to store this information and shall be segregated from
      other nonconfidential information.  Data generated from confidential
      samples shall be treated as confidential.  Upon receipt of
      confidential information,  the DCO logs these documents  into a
      Confidential Inventory Log.  The information is then made available
      to authorized personnel but only after it has been signed out to that
      person by the DCO.   The documents shall be  returned to  the locked
      file at the conclusion of each working day.   Confidential information
      may not be reproduced except upon approval  by the EPA Contracting
      Officer.   The DCO will enter all copies into the document control
      system.   In addition,  this information may  not be disposed of except
      upon approval by the EPA Contracting Officer.   The DCO  shall remove
      and retain the cover page  of any confidential information disposed of
      for one year and shall keep a record of the  disposition in the
      Confidential Inventory Log.
                                   F-6
Rev. 9/88

-------
                                  Figure 1
                                   Example
                             DOCUMENT INVENTORY
Document Control #*
                           Document Type
                                            # Pages
232-2-
232-2-
232-2-
232-2-
232-2-
232-2-
232-2-
232-2-
232-2-
232-2-
232-2
232-2
etc.
0001
0002
0003
0004
0005
0006
0007
0008
0009
0010
•0011
-0012
Case File Document Inventory Sheet
Chain-of-Custody Records
Shipping Manifests
Sample Tags
SMO Organics Traffic Reports
GC/MS spectra for sample B0310
GC/MS spectra for sample B0311
GC/MS spectra for sample B0319
Analyst's logbook pages
GC/MS library search worksheets
GC instrument log pages
GC/MS QC data sheets
etc.
   1
   2
   2
  50
  10
  20
  20
  20
   6
  15
   5
   4
etc.
*This number  is  to be  recorded  on  each  set of documents.

                                    F-7
                                                              Rev. 9/88

-------
    EXHIBIT G
GLOSSARY OF TERMS

-------

-------
                               GLOSSARY OF TERMS
ALIQUOT  -  a measured portion of a sample taken for analysis.

ANALYSIS DATE/TIME  - the date and military time of the injection of the
sample,  standard, or blank  into the GC/MS or GC system.

AROCLOR  -  trade name (Monsanto) for a series of commercial polychlorinated
biphenyls  and polychlorinated terphenyl mixtures marketed in the United
States.

BAR GRAPH  SPECTRUM  - a plot of the mass-to-charge ratio  (m/e) versus relative
intensity  of the  ion current.

BLANK  -  see Method  Blank

4-BROMOFLUOROBEN2ENE (BFB)  - compound chosen to establish mass spectral
tuning performance  for volatile analyses.  BFB tuning criteria must be met
before GC/MS analysis can begin.  (BFB is also used as a surrogate.)

CALIBRATION CHECK COMPOUNDS (CCC) - target compounds used to evaluate the
calibration stability (precision) of the GC/MS system.  Maximum percent
deviations of the CCCs are  defined in the protocol.

CASE - a finite, usually predetermined number of samples collected over a
given  time period from a particular site. Case numbers are assigned by the
Sample Management Office.   A case consists of one or more Sample Delivery
Groups.

CHARACTERIZATION  -  a determination of the approximate concentration range of
compounds of interest used  to choose the appropriate analytical protocol.

CONFIRMATION ANALYSIS - see Primary Analysis.

CONTINUING CALIBRATION - analytical standard run every 12 hours to verify the
calibration of the purge and trap-GC/MS system.

CONTINUOUS LIQUID-LIQUID EXTRACTION - used herein synonymously with the terms
continuous extraction,  continuous liquid extraction,  and liquid extraction.

CONTROL MATRIX SPIKE -  corn oil fortified with known quantities of specified
compounds analyzed  to measure the recovery (accuracy) of the entire
analytical method.

DAY - unless otherwise  specified,  day shall mean calendar day.

DECAFLUOROTRIPHENYLPHOSPHINE (DFTPP)  - compound chosen to establish mass
spectral tuning performance for extractable analysis.  DFTPP tuning criteria
must be met before GC/MS analysis can begin.

EXTRACTABLE - a compound that can be  partitioned into an organic solvent from
the sample matrix and is amenable to  gas  chromatography.   Extractables
include Base/Neutrals,  Acids and Festicide/PCB compounds.


                                   G-2                             Rev.  9/88

-------
HOMOLOG - one of the 10 levels of chlorination of PCBs (Cj^HgCl through '   -
^12CH(P or other SrouP °f compounds, varying by systematic addition of
substituent.

IN-HOUSE - at the Contractor's facility.

INITIAL CALIBRATION - analysis of analytical standards for a series of
different specified concentrations; used to define the.linearity and dynamic
range of the response of the mass spectrometer to the target compounds.

INTERNAL STANDARDS - compounds added to every standard, blank, control matrix.
spike and sample extract at a known concentration, prior to analysis.
Internal standards are used as the basis for quantitation of the target
compounds.

LABORATORY  - synonymous with Contractor as used herein.

HETHOD BLANK (previously termed reagent blank) - an analytical control
consisting  of all reagents, internal standards and surrogate standards, that
is carried  through the entire analytical procedure.  The method blank is used
to "define the level of laboratory background contamination.

NARRATIVE (Case Narrative) - first portion of the data package which includes
laboratory  name, contract number, Case and sample number identification, and
descriptive documentation of any problems encountered in processing the
samples, along with corrective action taken and problem resolution.  Complete
Case Narrative specifications are included in Exhibit B.

PERCENT MOISTURE - an approximation of the amount, of water in a solid phase
made by drying an aliquot of the sample at 105°C.  The percent moisture
determined  in this manner also includes contributions from all compounds that
may volatilize at 105°C, including water.  Percent moisture is determined
from decanted samples and from samples that are not decanted.

'PHASE - describes the physical state(s) 'of the sample.  Three "phase
designatores" are used:  Solid; Water miscible liquid; Water immiscible
liquid. . A  sample may contain multiple phases.

PCB CONGENER - one of 209 PCBs at any level of chlorination.

PCB ISOMER  - any PCB or other compound which has the same molecular formula,
but different positional substitutions.  (2,2' dichlorobiphenyl and 2,3
dichlorobiphenyl are isomeric.)

POLYCHLORINATED BIPHENYLS (PCBS) - a class of 209 discrete chemical compounds
in which one to ten chlorine atoms are attached to biphenyl; i.e.,
monochlorobiphenyl through decachlorobiphenyl with the formula Ci2^1
where n - 1-10.

PRIMARY ANALYSIS - one of two types of pesticide/PCB analysis by GC/EC
techniques, the other being the Confirmation Analysis.  If the two analyses
are run at  separate times, the Primary Analysis is the first analysis
                                    G-3                             Rev. 9/88

-------
chronologically, and is used to establish the tentative identification of any
pesticides/PCBs detected. The identification is then confirmed in the
confirmation analysis.  If the two analyses are simultaneous, either may be
considered  the Primary Analysis.

PROTOCOL  -  describes the exact procedures to be followed with respect to
sample receipt and handling, analytical methods, data reporting and
deliverables, and document control.  Used synonymously with Statement of Work
(SOW) .

PURGE AND TRAP (DEVICE) - analytical technique (device) used to isolate
volatile  (purgeable) organics by stripping the compounds from water or soil
by a stream of inert gas, .trapping the compounds on a porous polymer trap,
and thermally desorbing the trapped compounds onto the gas chromatographic
column.

REAGENT WATER - water in which an interferent is not observed at or above the
minimum quantitation limit of the parameters of interest.

RECONSTRUCTED ION CHROMATOGRAM (RIC) - a mass spectral graphical
representation of the separation achieved by a gas chromatograph ; a plot of
total ion current versus retention time.

RECOVERY  -  a determination of the accuracy of the analytical procedure made
by comparing measured values for a fortified (spiked) sample against the
known spike values.  Recovery is determined by the following equation:

%Rec -  measured value x 100%
         known value

RELATIVE RESPONSE FACTOR (RRF) - a measure of the relative mass spectral
response of an analyte compared to its internal standard.  Relative Response
Factors are determined by analysis of standards and are used in the
calculation of concentrations of analytes in samples.  RRF is determined by
the following equation:
RRF -
          A^   x    Cls
Where A  - area of the characteristic ion measured

      C  - concentration

      is - internal standard

      x  — analyte of interest

RESOLUTION - also termed separation, the separation between peaks on a
chromatograms ,  calculated by dividing the height of the valley between the
peaks by the peak height of the smaller peak being resolved, multipled by
100.
                                   G-4                              Rev.  9/88

-------
SAMPLE - a portion of material to be analyzed that is contained in single or
multiple containers and identified by a unique sample number.  A sample may
contain more than one phase.

SAMPLE DELIVERY GROUP (SDG) - a unit within a sample Case that is used to
identify a group of samples for delivery.  An SDG is a group of 20 or fewer
single phase units within a Case, received over a period of up to 14 calendar
days.  Data from all single phase units in an SDG are due concurrently.  A
Sample Delivery Group is defined by one of the following, whichever occurs
first:

     o    Case;  or

     o    Each 20 single phase units  within a Case;  or

     o    Each 14-day calendar period during which samples  in a Case are
          received,  beginning with receipt of the  first sample in the Case or
          SDG.

SAMPLE NUMBER (EPA Sample Number) - a unique identification number designated
by EPA for each sample.  The EPA sample number appears on the sample Traffic
Report which documents information on that sample.

SINGLE PHASE UNIT - a subsample consisting of a single phase.  Multiple phase
samples are phase separated into single phase units.  According to the
protocol, single phase units are analyzed and reported as discrete samples.

SOIL  - used herein synonymously with soil/sediment and sediment.

STANDARD ANALYSIS - an analytical determination made with known quantities of
target compounds; used to determine response factors and thereby calibrate
the mass spectrometer.

SURROGATES (Surrogate Standard)  - compounds added to every blank, sample,
control matrix spike, and standard; used to evaluate analytical efficiency by
measuring recovery.  Surrogates are brominated or isotopically labeled
compounds not expected to be detected in environmental media.

SYSTEM PERFORMANCE CHECK COMPOUNDS (SPCC) - target compounds designated to
monitor purge and trap/chromatographic performance, sensitivity and compound
instability or degradation on active sites. .Minimum response factor criteria
for the SPCCs are defined in the protocol.

TARGET COMPOUND LIST (TCL) - a list of compounds designated by the Statement
of Work (Exhibit C) for analysis.

TENTATIVELY IDENTIFIED COMPOUNDS  (TIC) - compounds detected in samples that
are not target compounds, internal standards or surrogate standards.  Up to
30 peaks (those greater than 10% of peak areas or heights of nearest internal
standards) are subjected to mass spectral library searches for tentative
identification.

TIME  - when required to record time on any deliverable item, time shall be
expressed as Military Time, i.e., a 24-hour clock.
                                    G-5
Rev. 9/88

-------
TRAFFIC REPORT (TR) - an EPA sample identification form filled out by the
sampler, which accompanies the sample during shipment to the laboratory and
which documents sample condition and receipt by the laboratory.

TWELVE-HOUR TIME PERIOD - The twelve (12) hour time period for GC/MS system
tuning and standards calibration (initial or continuing calibration) begins
at the moment of injection of the DFTPP or BFB analysis that the laboratory
submits as documentation of compliant tune.  The time period ends after 12
hours has elapsed according to the system clock.

VALIDATED TIME OF SAMPLE RECEIPT (VTSR) - the date on which a sample is
received at the Contractor's facility,  as recorded on the shipper's delivery
receipt and Sample Traffic Report.

VOLATILE COMPOUNDS - compounds amenable to analysis by the purge and trap
technique.  Used synonymously with purgeable compounds.

WIDE BORE CAPILLARY COLUMN - a gas chromatographic column with an internal
diameter (ID) that is greater than 0.32 mm.  Columns with lesser diameters
are classified as narrow bore capillaries.
                                   G-6
Rev. 9/88

-------

-------