EPA Document #: 815-R-05-007
METHOD 331.0    DETERMINATION OF PERCHLORATE IN DRINKING WATER BY
                 LIQUID CHROMATOGRAPHY ELECTROSPRAY
                 IONIZATION MASS SPECTROMETRY
                                   Revision 1.0
                                  January 2005
S.C. Wendelken and D. J. Munch (U.S. EPA, Office of Ground Water and Drinking Water)
B.V. Pepich (Shaw Environmental, Inc.)
D. W. Later and C. A. Pohl (Dionex, Inc. Sunnyvale, Ca)
                         TECHNICAL SUPPORT CENTER
               OFFICE OF GROUND WATER AND DRINKING WATER
                  U. S. ENVIRONMENTAL PROTECTION AGENCY
                             CINCINNATI, OHIO 45268
                                      331.0-1

-------
                                      METHOD 331.0

        DETERMINATION OF PERCHLORATE IN DRINKING WATER BY LIQUID
     CHROMATOGRAPHY ELECTROSPRAY IONIZATION MASS SPECTROMETRY
                                        (LC/ESI/MS)
1.  SCOPE AND APPLICATION

   1.1   This is a liquid chromatography electrospray ionization mass spectrometry (LC/ESI/MS)
         method for the determination of perchlorate in raw and finished drinking waters. This method
         can be used to acquire data using either Selected Ion Monitoring (SIM) or Multiple Reaction
         Monitoring (MRM) detection.  Based on known interferences, MRM detection is
         recommended, however, SIM detection may be used if all of the criteria outlined in Section
         9.1  are met. Precision and accuracy data have been generated for both SIM and MRM
         detection of perchl orate in reagent water, finished groundwater, finished surface water and a
         synthetic high ionic strength matrix. The single laboratory Lowest Concentration Minimum
         Reporting Level (LCMRL) has also been determined for both detection modes in reagent
         water.1
                                          Chemical Abstract Services
                           Analvte        Registry Number (CASRN)
                         Perchl orate              14797-73-0

   1.2   The SIM and MRM mass spectrometry conditions described in this method were developed
         using  a conventional LC system. Analysts interested in using an ion chromatography system
         with MS detection should consult EPA Method 332.0.

   1.3   The Minimum Reporting Level (MRL) is the lowest analyte concentration that meets the Data
         Quality Objectives (DQOs) which are based on the intended use of this method. The single
         laboratory LCMRL is the lowest true concentration for which the future recovery is predicted
         to fall between 50-150% recovery with 99% confidence.  Single laboratory LCMRLs for
         perchl orate were 0.022 ng/L for MRM using m/z 83 and 0.056 |J,g/L for SIM using m/z 101.
         These values are also provided in Table 2.  The procedure used to  determine the LCMRL is
         described elsewhere.1

   1.4   Laboratories using this method are not required to determine an LCMRL, but they will need
         to demonstrate that their laboratory MRL for this method meets the requirements described in
         Section 9.2.4.

   1.5   Detection limit (DL) is defined as the statistically calculated minimum concentration that can
         be measured with 99% confidence that the reported value is greater than zero2.  The DL is
         dependent on sample matrix, fortification concentration, and instrument performance.
         Determining the DL  for analytes in this method is optional (Sect. 9.2.5).  The DLs for
         perchlorate were 0.005 ng/L in LC/MRM mode and 0.008 ng/L in the LC/SIM Mode. These
         values are also provided in Table 2.

                                           331.0-2

-------
    1.6   This method is intended for use by analysts skilled in the operation of LC/MS instrumentation
         and the interpretation of the associated data.

2.  SUMMARY OF METHOD

    2.1    Water samples are collected in the field using a sterile filtration technique. Prior to analysis,
          isotopically enriched perchlorate (C118O4) is added to the sample as an internal standard.
          The sample is injected without cleanup or concentration onto a chromatographic column
          (Dionex lonPak® AS-21 or equivalent), which separates perchl orate from other anions and
          background interferences. Perchlorate is subsequently detected by negative electrospray
          ionization mass spectrometry.  A remote controlled valve is used to divert early eluting
          cations and anions to waste. Prior to the elution of perchl orate, the valve is switched sending
          the chromatographic eluent to the mass spectrometer.  This diversion helps prevent
          unnecessary fouling of the electrospray source. Perchlorate is quantified using the internal
          standard technique.

3   DEFINITIONS

    3.1   ANALYSIS BATCH - A sequence of samples, which are analyzed within a 30-hour period
         and include no more than 20 field samples. Each Analysis Batch must also include all
         required QC samples, which do not contribute to the maximum field sample total of 20. The
         required QC samples include:

         Laboratory Reagent Blank (LRB)
         Continuing Calibration Check Standards (CCCs)
         Laboratory Fortified Blank (LFB)
         Laboratory Fortified Sample Matrix (LFSM)
         Laboratory Fortified Sample Matrix Duplicate or Laboratory Duplicate (LFSMD or LD)

    3.2   CALIBRATION STANDARD  (C AL) - A solution of the target analyte prepared from the
         primary dilution standard solution or stock standard solution and the internal standard. The
         CAL solutions are used to calibrate the instrument response with respect to analyte
         concentration.

    3.3   CONTINUING CALIBRATION CHECK (CCC) - A calibration standard containing the
         method analyte, which is analyzed periodically to verify the accuracy of the existing
         calibration.

    3.4   DETECTION LIMIT (DL) - The minimum concentration of an analyte that can be identified,
         measured and reported with 99% confidence that the analyte concentration is greater than
         zero.  This is a statistical determination (Sect. 9.2.5), and accurate quantitation is not expected
         at this level.2

    3.5   INTERNAL STANDARD (IS)  - A pure compound added to all standard solutions and field
         samples in a known amount. It is used to measure the relative response of the method analyte.
         The internal standard must be a compound that is not a sample component.

                                            331.0-3

-------
3.6   LABORATORY DUPLICATES (LDs) - Two sample aliquots (LDi and LD2), taken in the
      laboratory from a single sample bottle, and analyzed separately with identical procedures.
      Analyses of LDi and LD2 indicate precision associated specifically with the laboratory
      procedures by removing variation contributed from sample collection, preservation and
      storage procedures.

3.7   LABORATORY FORTIFIED BLANK (LFB) - An aliquot of reagent water or other blank
      matrix to which a known quantity of the method analyte is added. The LFB is analyzed
      exactly like a sample including the preservation procedures in Section 8. Its purpose is to
      determine whether the methodology is in control, and whether the laboratory is capable of
      making accurate measurements.

3.8   LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - An aliquot of a field sample to
      which a known quantity of the method analyte is added.  The LFSM is processed and
      analyzed exactly like a sample, and its purpose is to determine whether the sample matrix
      contributes bias to the analytical results. The background concentration of the analyte in the
      sample matrix must be determined in a separate aliquot and the measured value in the LFSM
      corrected for background concentrations.

3.9   LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE (LFSMD) - A second
      aliquot of the field sample used  to prepare the LFSM which is fortified and analyzed
      identically to the LFSM.  The LFSMD is used instead of the Laboratory Duplicate to assess
      method precision and accuracy when the occurrence of the target analyte is infrequent.

3.10  LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX (LFSSM) - An aliquot of
      the Laboratory Synthetic Sample Matrix (Sect. 7.2.4) which is fortified with perchlorate.  This
      QC sample is processed like a field sample (Sect. 8.1) and is used to confirm that the analyst
      has adequate chromatographic resolution between sulfate and perchlorate.

3.11  LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or other blank
      matrix that is treated exactly as a sample including exposure to all filtration equipment,
      storage containers and internal standards. The LRB is used to determine if the method analyte
      or other interferences are present in the laboratory environment, the reagents, or the apparatus.

3.12  LABORATORY SYNTHETIC  SAMPLE MATRIX BLANK (LSSMB) - An aliquot of the
      Laboratory Synthetic Sample Matrix (Sect. 7.2.4) that is not fortified with perchlorate.  This
      QC sample is processed like a field sample (Sect. 8.1) and is used to determine if the method
      analyte or other interferences are present in the LSSM solution.

3.13  LOWEST CONCENTRATION MINIMUM REPORTING LEVEL (LCMRL) - The single
      laboratory LCMRL is the lowest true concentration for which the future recovery is predicted
      to fall between 50-150% recovery with 99% confidence.1

3.14  MATERIAL SAFETY DATA SHEETS (MSDS) - These sheets contain written information
      provided by vendors concerning a chemical's toxicity, health hazards, physical properties,
      fire, and reactivity data including storage, spill, and handling precautions.
                                       331.0-4

-------
   3.15  MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can be reported
         by a laboratory as a quantified value for the target analyte in a sample following analysis.
         This concentration must meet the criteria defined in Section 9.2.4 and must be no lower than
         the concentration of the lowest calibration standard for the target analyte.

   3.16  MULTIPLE REACTION MONITORING (MRM) - A mass spectrometric technique where a
         parent ion is first isolated then subsequently fragmented into a product ion(s).  Quantitation is
         accomplished by monitoring a specific product ion.

   3.17  PRIMARY DILUTION STANDARD SOLUTION (PDS) - A solution containing the method
         analyte prepared in the laboratory from stock standard solutions and diluted as needed to
         prepare Calibration Standards and other analyte solutions.

   3.18  QUALITY CONTROL SAMPLE (QCS) - A solution containing the method analyte at a
         known concentration which is obtained from a source external to the laboratory and different
         from the source of calibration standards. The QCS is used to verify the calibration
         standards/curve integrity.

   3.19  REAGENT WATER (RW) - Purified water which does not contain any measurable quantity
         of the target analyte or interfering compounds at or above 1/3 the MRL for the target analyte.

   3.20  SELECTED ION MONITORING (SIM) - A mass spectrometric technique where only select
         ions are monitored. This technique helps to increase sensitivity.

   3.21  STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing the method
         analyte that is prepared in the laboratory using assayed reference materials or purchased from
         a reputable commercial source.

4.  INTERFERENCES

   4.1   All glassware must be meticulously cleaned. Wash glassware thoroughly and rinse with
         reagent water.  As an alternative to glassware, disposable high-density polyethylene
         equipment may be used.

         NOTE: Perchl orate has been detected in some common laboratory detergents. If background
         contamination is a problem, detergents used during glassware cleaning should be investigated
         as a potential source of contamination.

   4.2   Method interferences may be caused by contaminants in solvents, reagents (including reagent
         water), sample bottles and caps, and other sample processing hardware. These interferences
         may lead to discrete artifacts and /or elevated baselines in the chromatograms. All laboratory
         reagents and equipment must be routinely demonstrated to be free from interferences (less
         than l/j the MRL for the target analyte) under the conditions of the analysis. This may be
         accomplished by analyzing LRBs as described in Section 3.11.  Subtracting blank values
         from  sample  results is not permitted.

                                           331.0-5

-------
   4.3   Matrix interferences may be caused by contaminants that are present in the sample.  The
         extent of matrix interferences will vary considerably from source to source, depending upon
         the nature of the water. Water samples high in organic carbon or dissolved solids may lead to
         elevated chromatographic baselines or interfering peaks.
       4.3.1  Hydrogen sulfate that is formed from the   S isotope of sulfur (H  SO4, m/z 99) is a
             common interference. It can interfere with the qualitative identification of perchlorate if
             there is poor chromatographic resolution between sulfate and perchlorate. This
             interference is particularly troublesome when using the SIM mode of detection, as the m/z
             of perchlorate (which is required for confirmation) is m/z 99. When using MRM
             detection, this ion, at modest concentrations, does not interfere with the identification or
             quantitation of perchlorate.

   4.4   Equipment used for sample collection and storage has the potential to introduce interferences.
         The potential for interferences from these devices must be investigated during the Initial
         Demonstration of Capability (Sect. 9.2) by preparing and analyzing a Laboratory Reagent
         Blank (LRB).  This procedure should be repeated each time that a new brand or lot of
         equipment is used to ensure that background contamination does not interfere with the
         identification and quantitation of perchlorate.

   4.5   The percent of 18O enrichment of the internal standard may vary between manufacturers.
         Poor isotopic enrichment of the internal standard may lead to sample contamination by
         C116O4 .  Consequently, it must be demonstrated that solutions containing a working level of
         the internal standard do not contain unlabeled perchlorate at concentrations greater than 1/3  of
         the MRL. This is initially confirmed during the analysis of the Laboratory Reagent Blank
         (LRB) during the IDC and is monitored in each Analysis Batch.

5.  SAFETY

   5.1   The toxicity or carcinogenicity of each reagent used in this method has not been precisely
         defined.  Each chemical should be treated as a potential health hazard, and exposure to these
         chemicals should be minimized. Each laboratory is responsible for maintaining an awareness
         of OSHA regulations regarding safe  handling of chemicals used in this method. A reference
         file of MSDSs should be made available to all personnel involved in the chemical analysis.
         Additional references to laboratory safety are available.3"5

6.  EQUIPMENT AND SUPPLIES (References to specific brands or catalog numbers are included as
         examples only and do not imply endorsement of the product.)

   6.1   SAMPLE CONTAINERS -  125-mL sterile high-density polyethylene (HOPE) bottles (I-
         Chem 125-mL sterile HOPE bottle, Fisher Cat. No. N411-0125 or equivalent).

   6.2   SAMPLE FILTERS - Sterile sample filters (Corning 26-mm surfactant free cellulose acetate
         0.2-um filter, Fisher Cat. No. 09-754-13 or equivalent).  If alternate filters are used they
         should be certified as having passed  a bacterial challenge test.6 Additionally, if alternate
         filters or different lots of the recommended filters are used, they must be tested using a LRB

                                            331.0-6

-------
      and a LFB fortified at the MRL as outlined in Section 9.2 to insure that they do not introduce
      interferences or retain perchlorate.

6.3    SYRINGES - Sterile, silicone free disposable syringes (Henke Sass Wolf 20-mL Luer lock,
      Fisher Cat. No.  14-817-33 or equivalent).

6.4    VOLUMETRIC FLASKS - Class A, suggested sizes include 5, 100, 250, 500 and 1000 mL
      for preparation of standards and mobile phase.

6.5    GRADUATED CYLINDER - Suggested size 25-mL.

6.6    AUTO PIPETTES - Capable of delivering variable volumes from 25 |iL to 2000 |iL.

6.7    ANALYTICAL BALANCE - Capable of weighing to the nearest 0.0001 g.

6.8    NITROGEN - High purity compressed gas used for desolvation in the mass spectrometer.
      The gas purity and pressure requirements will depend on the instrument manufacturers'
      specifications.

6.9    ARGON - High purity compressed gas used in the collision cell of the mass spectrometer.
      The gas purity and pressure requirements will depend on the instrument
      manufactures' specifications.

6.10  SAMPLE PRETREATMENT CARTRIDGES - Cartridges may be needed to clean up
      samples that can not be analyzed due to high levels of background interferences. Single-use,
      disposable cartridges (OnGuard-II Ba2+ Dionex, Cat. No.  57093 or equivalent) may be used to
      remove sulfate from the sample.

6.11  LIQUID CHROMATOGRAPHY ELECTROSPRAY MASS SPECTROMETRY SYSTEM
      (LC/ESI/MS)

   6.11.1 LC COLUMN - LC column, 2.1 x 250 mm (Dionex  IonPak®AS-21 or equivalent). Any
         column that provides  adequate resolution, peak shape, capacity, accuracy, and precision
         (Sect. 9.1) may be used.

         NOTE: The IonPak®AS-21 columns are shipped with a sodium hydroxide mobile phase.
         During method development new columns were flushed with 200 mM methylamine
         mobile phase for approximately 30 minutes before being connected to the mass
         spectrometer. Additionally, the retention time of perchlorate on a new lonPak® AS-21
         would decrease over several days as the column equilibrated.  This shift in retention time
         was inconsequential to the analysis due to the presence of the internal standard.

   6.11.2 LC SYSTEM - (Waters 2690 or equivalent) The LC  system must be compatible with a
         basic mobile phase (approximately pH 12). If the LC system contains Vespel®
         components, such as injection valve rotor seals, these components must be replaced with
         a material suitable for high use at high pH. Suitable materials for high pH include
         Tefzel®, PEEK™, Teflon®, UHMWPE and KEL-F®.  Consult the instrument
                                       331.0-7

-------
             manufacturer if there are questions regarding the use of a high pH mobile phase in the
             system.  Additionally, the system must be capable of consistently injecting up tolOO-|jL
             volumes and delivering mobile phase at a constant flow rate of 0.350 mL/min.

       6.11.3 MASS SPECTROMETER - (Micromass QuattroMicro or equivalent) The MS must be
             capable of providing electrospray ionization with negative ion detection. Due to the
             potential for H34SO4 interference, MRM detection is recommended.

       6.11.4 DATA SYSTEM - An interfaced data system is required to acquire, store,  and output MS
             data. The computer software should have the capability of processing stored LC/MS data
             by recognizing a chromatographic peak within a given retention time window. The
             software must allow integration of the ion abundance of any specific ion between
             specified time or scan number limits. The software must be able to construct a linear
             regression or quadratic calibration curve, and calculate analyte concentrations using the
             internal standard technique.

7.  REAGENTS AND STANDARDS

   7.1   REAGENTS AND SOLVENTS - Reagent grade or better chemicals should be used in all
         analyses. Unless otherwise indicated, it is intended that all reagents will conform to the
         specifications of the Committee on Analytical Reagents of the American Chemical Society
         (ACS), where such specifications are available. Other grades may be used, as long as the
         reagent is of sufficiently high purity to permit its use without lessening the quality of the
         determination.

       7.1.1   METHYLAMINE - (CASRN 74-89-5) - (40 wt.%) in water (Aldrich Cat. No. 426466 or
             equivalent).

       7.1.2   METHYLAMINE MOBILE PHASE (200 mM) - Add 20 mL of the 40 wt.%
             methylamine solution in water to a 1L volumetric flask.  Dilute to volume with reagent
             water. During method development, the mobile phase was found to be stable for at least
             48 hours. Laboratories are encouraged to institute their own quality control (QC)
             procedures to determine when solutions need to be replaced.

       7.1.3   REAGENT WATER - Purified water that does not contain any measurable quantity of
             the target analyte or interfering compounds at or above V3 the MRL for the target analyte
             (Sect. 3.19). The reagent water used during method development was generated from tap
             water using a Millipore ELIX-3 followed by a Millipore Gradient A10 system.

       7.1.4   SODIUM CHLORIDE - (NaCl, CASRN 7647-14-5) - Fisher Cat. No. S-271 or
             equivalent.

       7.1.5   SODIUM BICARBONATE - (NaHCO3, CASRN 497-19-8) - Fluka Cat. No. 71627 or
             equivalent.

       7.1.6   SODIUM SULFATE - (Na2SO4, CASRN 7757-82-6) - Fluka Cat. No. 71959 or
             equivalent.
                                           331.0-8

-------
7.2   STANDARD SOLUTIONS - When a compounds' purity is assayed to be 96 percent or
      greater, the weight can be used without correction to calculate the concentration of the stock
      standard.  Solution concentrations listed in this section were used to develop this method and
      are included as an example.  Stock standard solutions are estimated to be stable for one year
      and any fortified or dilute solutions made from the stock standards are stable for at least 30
      days  Although estimated stability times for standard solutions are given, laboratories
      should use standard QC practices to determine when their standards need to be
      replaced.

   7.2.1  INTERNAL STANDARD SOLUTION - This method uses the internal standard
         NaCl18O4.

        7.2.1.1   INTERNAL STANDARD STOCK STANDARD (ISSS)   (100 ng/mL C118O4) -
                Prepare, or purchase commercially, an ISSS at a concentration of 100 |j,g /mL.  To
                prepare this solution from a solid  standard, weigh out 12.1 mg of NaCl18O4into a
                100-mL volumetric flask and dilute to 100 mL with reagent water.  The internal
                standard used during method development was custom synthesized by Isotech
                (Miamisburg, OH).  Subsequently, several manufacturers offer this product as a
                standard item. Because the degree of 18O enrichment could vary between
                manufacturers, it must be confirmed during the IDC (Sect. 9.2) and in subsequent
                LRBs that this standard does not contribute unenriched perchlorate to the sample.
                Any background contribution from the IS must be < 1/3 of the perchlorate MRL
                when added at working levels (the concentration of the internal  standard used
                during method development was 1 ng/mL).

        7.2.1.2   INTERNAL STANDARD FORTIFICATION SOLUTION (77.0 ng/mL C118O4) -
                Place 77.0 |jL of the internal standard ISSS into a 100-mL volumetric flask and
                dilute to volume with reagent water.  The addition of 25 |jL to 1.9 mL of sample
                will produce a final concentration of 1.0 ng/mL.  All field samples and CAL
                standards must have the same amount of internal standard added.

   7.2.2  PERCHLORATE STANDARD SOLUTIONS - Obtain the analyte as a solid standard of
         NaClO4 or as a commercially prepared solution from a reputable standard manufacturer.
         Prepare the perchlorate stock, primary and secondary standards as described below.

       7.2.2.1   PERCHLORATE STOCK STANDARD SOLUTION (PSSS) (100 ng/mL C1O4 ) -
               To prepare this solution from a solid NaClO4 standard, weigh out  12.3 mg of
               NaClO/tinto a  100-mL volumetric flask and dilute to volume with reagent water.

       7.2.2.2   PERCHLORATE PRIMARY DILUTION SOLUTION (PPDS) (1.0 |ig/mL C1O4 )
               - Prepare the perchlorate PDS by adding 1.0 mL of the Stock Standard Solution to a
               100-mL volumetric flask and dilute to volume with reagent water. This solution is
               used to prepare the Perchlorate Secondary Dilution Standard and the Calibration
               Standards below.

                                       331.0-9

-------
   7.2.2.3  PERCHLORATE SECONDARY DILUTION SOLUTION (PSDS) (5.0 ng/mL
            C1O4 ) - Prepare the 5 ng/mL Analyte SDS by adding 500 |iL of the Analyte PDS
           to a 100-mL volumetric flask and dilute to volume with reagent water. This
           solution is used to prepare the Calibration Standards below.

   7.2.2.4  PERCHLORATE FORTIFICATION SOLUTIONS (PFS) (100, 200 and 1000
           |ig/mL) - The Analyte Fortification Solutions are prepared by dilution of the
           Perchlorate Secondary Dilution Solution and are used to fortify the LFB, LFSSM,
           the LFSMs and the LFSMDs with method analytes.  It is recommended that
           multiple concentrations be prepared so that the fortification levels can be rotated or
           adjusted to the concentration of the target analyte in the native samples.

7.2.3   CALIBRATION (CAL) STANDARDS - Prepare a calibration curve from dilutions of
       the perchlorate PDS and the perchlorate SDS using a minimum of five Calibration
       Standards that span the concentration range of interest. The lowest concentration
       calibration standard must be at or below the MRL. An example of the dilutions used to
       prepare the Calibration Standards used to collect the data in Section  17 are shown in the
       table below.
PREPARATION OF CALIBRATION CURVE STANDARDS
CAL
Level
1
2
3
4
5
6
Vol. of
Analyte PDS
(HL)



50
100
500
Vol. of
Analyte SDS
(HL)
400
1000
2000



Final Vol. of CAL
Std. (mL)
100
100
100
100
100
100
Final Cone, of CAL
Std. (ng/mL)
0.020
0.050
0.10
0.50
1.0
5.0
       A constant amount of the internal standard is added to each Calibration Standard (1.0
       ng/mL in the final volume).  This is accomplished for each standard by taking 1.9 mL of
       the calibration standard and placing it in a 2.0-mL autosampler vial and adding 25 |jL of
       the 77 ng/mL Internal Standard Fortification Solution.

       NOTE: CAL standards are not processed with the sample collection equipment.  This
       step must be omitted for the CALs in order to identify any potential losses associated
       with the sample filtration or collection protocols.
7.2.4  LABORATORY SYNTHETIC SAMPLE MATRIX (LSSM) - Prepare a LSSM that
      contains the common anions chloride, sulfate and bicarbonate at 1000 mg/L follows.

   7.2.4.1  Weigh out 1.40 g of NaHCO3, 1.48 g of Na2SO4, and 1.54 g of NaCl (Fluka 71627,
           Fluka 71959, Fisher S-271, respectively or equivalent).  Add these to a 1-L
           volumetric flask using a powder funnel and dilute to volume using reagent water.
                                    331.0-10

-------
       7.2.5  LABORATORY FORTIFIED SYNTHETIC SAMPLE MATRIX (LFSSM) - An aliquot
             of the Laboratory Synthetic Sample Matrix fortified using the Perchlorate Fortification
             Solutions prepared above (Sect. 7.2.2.4).

8.  SAMPLE COLLECTION, PRESERVATION, AND STORAGE

   8.1   SAMPLE COLLECTION

       8.1.1  Grab samples must be collected in accordance with conventional sampling practices.7

       8.1.2  When sampling from a cold water tap, open the tap and allow the system to flush until the
             water temperature has stabilized (usually approximately 3 to 5 minutes). Collect a
             representative sample from the flowing system using a beaker of appropriate size.  Use
             this bulk sample to generate individual samples as needed. A volume of at least 20-mL is
             required for each individual sample.

       8.1.3  When sampling from an open body of water, fill a beaker with water sampled from a
             representative area. Use this bulk  sample to generate individual samples as needed. A
             volume of at least 20-mL is required for each individual field sample.

       8.1.4  Once representative  samples are obtained, they must be filtered to remove any native
             microorganisms. Perchlorate is known to be susceptible to microbiological degradation
             by anaerobic bacteria8. Samples are filtered to remove microbes and stored with
             headspace to minimize the possibility that anaerobic  conditions develop during storage.
             At a minimum, leave the top one third of the bottle empty to reduce the potential for
             degradation by any remaining anaerobic organisms.

          8.1.4.1 Remove a sterile syringe (Sect. 6.3) from its package and draw up 20 mL of the bulk
                sample. Remove a sterile syringe filter (Sect 6.2) from its package without touching
                the exit Luer connection.  Connect the filter to the syringe making sure that no water
                from the syringe drops on the  exterior of the filter. Open a sterile sample container
                (Sect. 6.1) without touching the interior. Using gentle pressure, pass the sample
                through the filter into the sample container. During this process do not let the syringe
                or filter make contact with the sample container.  Following filtration, seal the  sample
                container tightly, label and prepare the container for shipment. Syringes and filters
                are single use items and must be discarded after each sample.

   8.2   SAMPLE SHIPMENT AND  STORAGE - Samples must be chilled during shipment and
          must not exceed 10 °C during the first 48 hours after collection.  Samples should be
          confirmed to be at or below 10 °C when they are received at the laboratory. Samples stored
          in the lab must be held at or below 6 °C until analysis.  Samples should not be frozen.
                                           331.0-11

-------
   8.3    SAMPLE HOLDING TIMES - Samples should be analyzed as soon as possible. Samples
          that are collected and stored as described in Sections 8.1 and 8.2 may be held for a maximum
          of 28 days.

9.  QUALITY CONTROL

   9.1   Quality Control (QC) requirements include the Initial Demonstration of Capability and
         ongoing QC requirements that must be met when preparing and analyzing field samples. This
         section describes each QC parameter, its required frequency, and the performance criteria that
         must be met in order to meet EPA quality objectives. The QC criteria discussed in the
         following sections are summarized in Section 17, Tables 5 and 6. These QC requirements are
         considered the minimum acceptable QC criteria. Laboratories are encouraged to institute
         additional QC practices to meet their specific needs.

       9.1.1  METHOD MODIFICATIONS - The analyst is permitted to modify the separation
             technique, LC columns, mobile phase composition, LC conditions and MS conditions.
             However, each time such method  modifications are made, the analyst must repeat the
             procedures of the IDC (Sect. 9.2).

   9.2   INITIAL DEMONSTRATION OF CAPABILITY (IDC) - The IDC must be successfully
         performed prior to analyzing any field samples. Prior to conducting the IDC, the analyst must
         meet the calibration requirements outlined in Section 10. Requirements for the initial
         demonstration of laboratory capability are described in the following sections and are
         summarized in Table 5.

       9.2.1  DEMONSTRATION OF LOW SYSTEM BACKGROUND - Analyze a LRB  and a
             LSSMB after processing both through all sample collection steps outlined in Section 8.1.
             Confirm that the blanks are reasonably free of contamination and that the criteria in
             Section 9.3.1 are met.  The LRB and LSSMB must be filtered using the same sample
             collection devices that are used for field samples.

          9.2.1.1  Concentration dependent carry-over is manifest by signals in subsequent samples
                  that increase proportionally to the concentration of the previously injected standard.
                  Analysis of reagent water blank (non-filtered)  must be performed after the highest
                  CAL standard during the IDC to determine if carry-over is present. The results for
                  this sample must meet the criteria outlined in section 9.3.1.  If the reagent water
                  blank fails to meet the criteria there is likely system carry-over.  The  source can
                  often be traced to injection valve problems  or an inadequate autosampler rinse
                  protocol. System carry-over should be eliminated by determining the source of the
                  problem and taking corrective action.

       9.2.2  DEMONSTRATION OF PRECISION - Prepare and analyze 7 replicate LFBs and
             LFSSMs.  These samples should be fortified near the midrange of the initial  calibration
             curve. All samples must be processed using the  sample collection devices described in
             Section 8.1. The percent relative standard deviation (%RSD) of the concentrations of the
             replicate analyses must be < 20%.
                                           331.0-12

-------
            n / ^ o^  Standard Deviation of Measured Concentrations  , nn
            %RSD =	xlOO
                                Average Concentration

9.2.3   DEMONSTRATION OF ACCURACY - Using the same sets of replicate data generated
       for section 9.2.2, calculate the average percent recovery. The average percent recovery
       of the replicate analyses must be within 80-120% of the true value.

                .. _.          Average Measured Concentration  , ^
                % Recovery =	x 100
                                 Fortified Concentration

9.2.4   MINIMUM REPORTING LEVEL (MRL) CONFIRMATION - Establish a target
       concentration for the MRL based on the intended use of the method.  The lowest
       calibration standard used to establish the initial calibration (as well as the low-level
       Continuing Calibration Check) must be at or below the concentration of the MRL.
       Establishing the MRL concentration too low may cause repeated failure of ongoing QC
       requirements. Confirm or validate the MRL following the procedure outlined below.

   9.2.4.1  Fortify and analyze seven replicate Laboratory Fortified Blanks at the target MRL
           concentration. All samples must be processed using the sample collection devices
           described in Section 8.1. Calculate the mean (Mean) and standard deviation (S) for
           these replicates. Determine the Half Range for the prediction interval of results
           (HRpiR) using the equation below

                                 HRpiR = 3.9638

            where S is the standard deviation, and 3.963 is a constant value for seven
            replicates.1

   9.2.4.2  Confirm that the upper and lower  limits for the Prediction Interval of Result (PIR =
           Mean + HRPiR) meet the upper and lower recovery limits as shown below

           The Upper PIR Limit must be <150 percent recovery.

                                   Mean + HRPR	xlOO<150%
                               Fortified Concentration

            The Lower PIR Limit must be > 50 percent recovery.

                                  Mean - HR
                                            "PIR
                              Fortified Concentration
                                                     100>50%
   9.2.4.3  The MRL is validated if both the Upper and Lower PIR Limits meet the criteria
           described above. If these criteria are not met, the MRL has been set too low and
           must be determined again at a higher concentration.

                                     331.0-13

-------
   9.2.5  DETECTION LIMIT DETERMINATION (optional) - While DL determination is not a
          specific requirement of this method, it may be required by various regulatory bodies
          associated with compliance monitoring. It is the responsibility of the laboratory to
          determine ifDL determination is required based upon the intended use of the data.

          Analyses for this procedure should be done over at least 3 days. Prepare at least 7
          replicate LFBs processing the reagent water through all sample collection steps outlined
          in Section 8.1. Use the solutions described in Section 7.2.2.4 to fortify at a concentration
          estimated to be near the DL.  This concentration may be estimated by selecting a
          concentration at 2-5 times the noise level. Analyze the seven replicates through all steps
          of Section 11.

          NOTE: If an MRL confirmation data set meets these requirements, a DL may be
          calculated from the MRL confirmation data, and no additional  analyses are necessary.

          Calculate the DL using the following equation:

                       DL= St(n_iji_a = o.99)

                 where:
                 t(n-i,i-a = 0.99)= Student's t value for the 99%  confidence level with n-1 degrees of
                               freedom
                 n = number of replicates
                 S = standard deviation of replicate analyses.

          NOTE: Do not subtract blank values when performing DL calculations.

9.3    ONGOING QC REQUIREMENTS - This section describes the ongoing QC criteria that must
      be followed when processing and analyzing field samples. Table 6 summarizes these
      requirements.

   9.3.1  LABORATORY REAGENT BLANK (LRB) -  A LRB is analyzed during the IDC and is
          required with each Analysis Batch (Sect. 3.1) to confirm that background contaminants
          are not interfering with the identification or quantitation of perchlorate. If the LRB
          produces a peak within the perchlorate retention time window, identify the source of
          contamination and eliminate the interference before processing samples. The LRB must
          contain the IS at the same concentration used to  fortify all field samples and CAL
          standards. LRBs must be processed in the same  manner as field samples including
          exposure to all sample collection devices (i.e., sterile filtration  into sample containers). If
          samples are collected using devices that have not been previously evaluated by the
          laboratory, duplicates of the sample collection devices must be sent with the samples so a
          LRB (and a LFB) may be processed in the laboratory.  If perchlorate or other
          interferences are detected in the LRB at concentrations > 1/3 of the MRL, then all data
          for the analyte must be considered invalid for all samples in the Analysis Batch.
                                        331.0-14

-------
       NOTE: Although quantitative data below the MRL may not be accurate enough for data
       reporting, such data are useful in determining the magnitude of a background
       interference. Therefore, blank contamination levels may be estimated by extrapolation,
       when the concentration is below the lowest calibration standard.

 9.3.2  CONTINUING CALIBRATION CHECK (CCC) - CCC standards are analyzed at the
       beginning of each Analysis Batch, after every ten field samples and at the end of the
       Analysis Batch. See Section 10.4 for concentration requirements and acceptance criteria.
 9.3.3  LABORATORY FORTIFIED BLANK (LFB) - A LFB is required with each Analysis
       Batch. The LFB fortification level must be rotated between low, medium, and high from
       batch to batch. The low concentration LFB must be as near as practical to the MRL.
       Results of LFBs fortified at < MRL must be within 50-150% of the true value. Results of
       LFB analysis from all other concentrations must be 80-120% of the true value. If the LFB
       results do not meet these criteria, then all data for perchlorate must be considered invalid
       for all samples in the Analysis Batch.

       NOTE: LFBs must be processed in the same manner as field samples including exposure
       to all sample collection devices (i.e., sterile filtration into sample containers). If samples
       are collected using devices that have not been previously evaluated by the laboratory,
       duplicates of the sample collection devices must be sent with the samples so a LFB (and a
       LRB) may be processed in the laboratory.

 9.3.4  INTERNAL STANDARD (IS) - The analyst must monitor the peak area of the internal
       standard in all injections  of the Analysis Batch. The IS response (as indicated by peak
       area) for any chromatographic run must not deviate by more than + 30% of the IS area
       measured during the first CCC of the Analysis Batch. If the IS area counts for a sample
       do not meet this criterion, inject a second aliquot of the suspect sample as part of the
       same or a new Analysis Batch.

   9.3.4.1   If the reinjected aliquot produces an acceptable internal standard response, report
            results for that aliquot.

   9.3.4.2   If the IS area counts of the reinjected aliquot still do not meet the IS criterion, check
            the IS area of the most recent CCC.  If the IS criterion is met in the CCC but not the
             sample, report the sample results as suspect/matrix.

   9.3.4.3   If the IS area criterion is not met in both the sample and the CCC, instrument
            maintenance such as sample cone cleaning may be necessary. Perform the
            appropriate instrument maintenance and then reinject the sample in a subsequent
            analytical batch.

9.3.5   ISOTOPE AREA COUNT RATIO ACCEPTANCE CRITERIA - All CAL standards,
       QC samples and field samples must meet the area count ratio requirement for
       confirmation of perchlorate.  Perchlorate has a strong M+2 ion due to the presence of the
       37C1 isotope. The analyst must confirm that the ratio of the molecular ion to its M+2 ion
                                     331.0-15

-------
       (99/101) for the SIM technique, or the ratio of their product ions (83/85) for the MRM
       technique, are within ± 25% of the theoretical value of 3.08 (2.31 to 3.85)9. If a CAL
       standard or QC sample fails this requirement, remedial action is required.  Necessary
       actions may include mass calibration, column replacement/regeneration or instrument
       cleaning. If the area count ratio for a field sample falls outside this range, the cause is
       frequently interference from sulfate in the sample matrix. In this case the sample may
       require dilution or pretreatment with a barium cartridge to remove the sulfate (see Sect
       11.3.4.2 regarding required remedial action).

9.3.6   RELATIVE RETENTION TIME ACCEPTANCE CRITERIA - All CAL standards, QC
       samples and field samples must meet the Relative Retention Time (RRT) requirement for
       confirmation of perchlorate. The RRT of perchlorate can be calculated using the
       equation

                      Retention Time of Perchl orate Ion
                                                         RRI
                   Retention Time of Internal Standard Ion

       The RRT must be 1.0 + 2% (0.98-1.02) for a peak to be identified as perchlorate.

 9.3.7  LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - Analysis of a LFSM (Sect.
       3.8) is required in each Analysis Batch. The LFSM is used to  assess whether the sample
       matrix affects method  accuracy. If a variety of different sample matrices are analyzed
       regularly, for example drinking water from groundwater and surface water sources,
       performance data should be collected for each source.

    9.3.7.1  Within each Analysis Batch, a minimum of one field sample is fortified as an LFSM
            for every 20 samples analyzed. The LFSM is prepared by fortifying a sample with
            an appropriate amount of the Perchlorate Fortification Solution (Sect. 7.2.2.4). The
            fortification  should be delivered in the smallest volume possible to minimize
            dilution of the sample. Select a spiking concentration  that is greater than or equal to
            the native background concentration, if known. Selecting a duplicate aliquot of a
            sample that has already been analyzed aids in the  selection of an appropriate spiking
            level. If this  is not possible, use historical data and rotate through low, medium and
            high calibration concentrations when selecting a fortifying concentration.

    9.3.7.2  Calculate the perchlorate recovery (%REC) using the equation
            A = measured concentration in the fortified sample
            B = measured concentration in the unfortified sample
            C = fortification concentration.

    9.3.7.3  Recoveries for samples fortified at concentrations < MRL should be 50-150%.
            Recoveries for samples fortified at all other concentrations should be 80-120%.  If

                                     331.0-16

-------
           the accuracy for a LFSM falls outside the designated range, and the laboratory
           performance for perchlorate is shown to be in control in the CCCs, the recovery is
           judged to be matrix biased. The result for perchlorate in the unfortified sample is
           labeled suspect/matrix to inform the data user that the results are suspect due to
           matrix effects.

       9.3.7.3.1  Field Samples that have native perchlorate concentrations below the MRL and
                are fortified at concentrations at or near the MRL should be corrected for the
                native levels in order to obtain meaningful %REC values. This is the only
                case where background subtraction of results below the MRL is
                permitted.

9.3.8   LABORATORY DUPLICATE OR LABORATORY FORTIFIED SAMPLE MATRIX
       DUPLICATE (LD or LFSMD) - Within each Analysis Batch, a minimum of one
       Laboratory Duplicate (LD) or Laboratory Fortified Sample Matrix Duplicate (LFSMD)
       must be analyzed.  Laboratory Duplicates check the precision associated with laboratory
       procedures.  If perchlorate is not routinely observed in field samples, a LFSMD should be
       analyzed rather than a LD.

   9.3.8.1  Calculate the relative percent difference (RPD) for duplicate measurements (LDi
           and LD2) using the equation

                                   ILD, -LD,
                           RPD       l
                                  (LDj+LDj/2

  9.3.8.2   RPDs for Laboratory Duplicates should be < 20%. Greater variability may be
           observed when Laboratory Duplicates have perchlorate concentrations that are
           within a factor of < 2 x MRL.  At these concentrations, Laboratory Duplicates
           should have RPDs that are < 50%.  If the RPD of the analyte falls outside the
           designated range, and the laboratory performance for the analyte is shown to be in
           control in the CCC, the precision is judged to be matrix influenced.  The result from
           the unfortified sample  is labeled suspect/matrix to inform the data user that the
           results are suspect due to matrix effects.

  9.3.8.3   If a LFSMD is analyzed instead of a Laboratory Duplicate, calculate the relative
           percent difference (RPD) for duplicate LFSMs (LFSM and LFSMD) using the
           equation

                                 LFSM - LFSMD
                       RPD=-,	r—xlOO
                               (LFSM

  9.3.8.4   RPDs for duplicate LFSMs should be < 20%. Greater variability may be observed
           when fortified LFSMs have analyte concentrations that are within < 2 x MRL.
           LFSMs at these concentrations should have RPDs that are < 50%. If the RPD of the
           analyte falls outside the designated range, and the laboratory performance for the
                                    331.0-17

-------
                  analyte is shown to be in control in the CCC, the precision is judged to be matrix
                  influenced. The result from the unfortified sample is labeled suspect/matrix to
                  inform the data user that the results are suspect due to matrix effects.

9.4    QUARTERLY INSTRUMENT PERFORMANCE CHECK USING THE LSSMB AND LFSSM
       - Analysis of a LFSSM (Sect. 3.10) must be performed at least quarterly to assess instrumental
       performance with respect to samples  high in common anions.  An aliquot of LFSSM (fortified at
       the mid-range of the calibration curve) must be processed and analyzed as a sample along with a
       LSSMB. Both solutions must be from the same stock of LSSM. Results for the LSSMB and
       LFSSM should meet the criteria set forth in Sections 9.3.1 and 9.3.3, respectively. If the LSSMB
       contains perchlorate at a concentration > 1/3 the MRL, then the source of the contamination
       should be identified and eliminated.  If the LFSSM does not meet the QC acceptance criteria for
       LFB recovery, instrument maintenance or column replacement may be required.

10. CALIBRATION AND STANDARDIZATION

    10.1   Demonstration and documentation of acceptable MS mass calibration and initial analyte
          calibration are required before any samples are analyzed. If the initial calibration is
          successful, continuing calibration checks are required at the beginning and end of each period
          in which analyses are performed,  and after every tenth sample. Verification of mass
          spectrometer mass calibration should be repeated each time a major instrument modification
          is made or maintenance is performed.

          Note: CAL solutions and CCC standards are not processed with the sample collection
          devices.  This step must be  omitted for the CALs and CCCs in order to identify any potential
          losses associated with the sample filtration or collection devices._

    10.2   MASS SPECTROMETER CALIBRATION AND OPTIMIZATION

       10.2.1 MASS CALIBRATION - Calibrate the mass spectrometer according to the
             manufacturer's recommendations. The user should be aware that many current LC/MS
             instruments are designed to analyze macromolecules having large m/z ratios.
             Subsequently, many of the LC/MS calibration procedures are designed to cover the full
             scanning range of the instrument. Since this method uses the lower portion of the mass
             range, it may be necessary to  use alternate calibration compounds that provide ions of a
             lower m/z ratio to properly define the mass calibration over the range needed for the
             analysis of perchl orate.  For our studies, a sodium iodide/rubidium iodide mixture was
             used as a calibration compound.  It is recommended that the analyst contact the
             instrument manufacturer regarding appropriate mass calibration standards. Mass
             calibration must be verified prior to the analysis of field samples by injecting a high-level
             calibration standard and acquiring a full scan continuum mass spectrum that covers the
             m/z range  of the target analyte. The perchl orate ion (35 C1O4) should have a m/z of 99 ±
             0.3, the (37C1O4) isotope should have a m/z of 101 ± 0.3 and the internal standard
             (35 Cl 18O4) should have a m/z of 107 ± 0.3.

       10.2.2 OPTIMIZING MS PARAMETERS - LC/MS instruments have a large number of
                                           331.0-18

-------
          instrumental parameters that must be optimized. Each LC/MS system will have different
          optimal conditions which are influenced by the source geometry and system design. Due
          to the differences in design, the recommendations of the instrument manufacturer should
          be followed when tuning the instrument. MS conditions should be established by
          infusing a solution of the target compound into the mobile phase while the analyst tunes
          (or varies) the MS parameters using the same mobile phase flow rate and composition as
          in the final analysis. The response for the parent ions must be optimized for MS analysis
          and three MS/MS transitions must be optimized for the MRM analysis of perchlorate.
          These transitions are listed in the following table.
Precursor Ion (m/z)
35C1O4(99)
37C1O4(101)
35C118O4(107)
Fragment Lost (m/z )
16O (16)
16O (16)
18O(18)
Product Ion (m/z )
35C103(83)
37C1O3 (85)
35C118O3(89)
   10.2.3  INSTRUMENT CONDITIONS - Operating conditions are described in Section 17
          Table 1. Conditions different from those described may be used if the QC criteria in
          Section 9.2 are met.  Different conditions include alternate LC columns, mobile phases
          and MS conditions.

10.3   INITIAL CALIBRATION - During method development daily calibrations were performed,
      however, it is permissible to verify the calibration with daily CCCs. Calibration must be
      performed using peak areas and the internal standard technique.  Calibration using peak
      heights is not permitted.

   10.3.1  CALIBRATION STANDARDS - Prepare a set of at least five CAL standards as
          described in Section 7.2.3. The lowest concentration of the calibration standards must be
          at or below the MRL. The MRL must be confirmed using the procedure outlined in
          Section 9.2.4 after establishing the initial calibration. Additionally, field samples must be
          quantified using a calibration curve that spans the same concentration range used to
          collect the IDC data (Sect. 9.2).

   10.3.2  CALIBRATION - The LC/MS system is calibrated using peak areas and the internal
          standard technique.  Concentrations may be calculated through the use of a linear or
          quadratic calibration curve. Quantitation ions are listed in Table 1 for both the MRM and
          SIM detection techniques.

   10.3.3  CALIBRATION ACCEPTANCE CRITERIA - The validation of the calibration is
          determined by calculating the concentration of the analyte from each of the analyses used
          to generate the calibration curve. Calibration points that are < MRL should calculate to be
          50-150% of their true value. All other calibration points should calculate to be 80 to
          120% of their true value. If these criteria cannot be met, the  analyst will have difficulty
          meeting ongoing QC criteria. In this case,  corrective action should be taken to reanalyze
          the calibration standards, restrict the range  of calibration, or select an alternate method of
          calibration.
                                        331.0-19

-------
       10.3.4 INITIAL CALIBRATION VERIFICATION - Analyze a QCS sample fortified near the
             midpoint of the calibration range. The QCS sample should be from a source different
             than the source of the calibration standards. If a second vendor is not available, then a
             different lot of the standard should be used. The QCS should be prepared and analyzed
             just like a CCC. The acceptance criterion for the QCS is that the calculated amount of
             perchlorate must be 80-120% of the true value. If the measured analyte concentration
             does not meet this criterion, check the entire analytical procedure to locate and correct the
             problem before analyzing any field samples.

   10.4   CONTINUING CALIBRATION CHECKS (CCCs) - The CCC verifies the calibration at the
          beginning, after every tenth field sample and at the end of each Analysis Batch. CCCs are
          not counted as samples. The beginning CCCs for each Analysis Batch must include a
          CCC at or below the MRL and a CCC fortified at a level  near the midpoint of the
          curve.  These CCCs verify instrument sensitivity and the accuracy of the curve prior to the
          analyses of samples. Subsequent CCCs should alternate between a low, medium and high
          concentration.

       10.4.1 Inject an aliquot of the appropriate concentration calibration solution and analyze with
             the same conditions used during the initial calibration.

       10.4.2 Determine that the absolute area of the quantitation ion of the internal standard has not
             changed by more than + 30% from the first CCC of the Analysis Batch.

             NOTE: The IS area counts of the first CCC of the Analysis Batch should be > 50% of
             the average of the IS area counts of the CAL standards from the initial calibration.  If the
             IS response drifts below 50% of the average determined during the initial calibration,
             instrument maintenance or ESI/MS detector inlet cleaning is generally required.

       10.4.3 Calculate the concentration of the analyte in the CCC.  CCCs fortified at < MRL must
             calculate to be 50-150% of the true value.  CCCs fortified at all other levels must
             calculate to be 80-120%.  If these conditions do not exist, then all data for the analyte
             must be considered invalid, and remedial action (Sect. 10.4.4) should be taken. The
             remedial action may require recalibration.  Any field samples that have been analyzed
             since the last  acceptable calibration verification and are still within holding time should
             be reanalyzed after adequate calibration has been restored.

       10.4.4 REMEDIAL  ACTION - Failure to meet CCC QC performance criteria requires remedial
             action. Maintenance such as cleaning an ion source probe,  sample cone, ion lenses
             and/or regenerating  or replacing LC columns may be required. Following major
             maintenance, the analyst must return to the initial calibration step (Sect. 10.3).

11. PROCEDURE

   11.1   Important aspects of this analytical procedure include proper sample collection and storage
          (Section 8), ensuring that the instrument is properly calibrated (Section 10) and that all
          required QC are met (Section 9). This section describes the procedures for sample
          preparation and analysis.
                                           331.0-20

-------
11.2   SAMPLE PREPARATION

    11.2.1  Collect and store field samples as described in Section 8.1. Allow field samples to reach
          room temperature prior to analysis.

    11.2.2  Process all LRBs, LFBs, LFSSMs and LFSSMBs using the sample collection devices
          described in Section 8.1.

    11.2.3  Transfer a 1.9-mL aliqout of each Field or QC Sample to an autosampler vial. Add 25
          |jL of the IS Fortification Solution (Sect. 7.2.1.2), cap and mix well. Larger amounts of
          sample may be used, but the amount of internal standard must be increased to give a final
          concentration that is equal to the CAL IS concentration.

11.3   SAMPLE ANALYSIS

    11.3.1  Establish operating conditions as described in Table 1 of Section 17.

    11.3.2  Establish a valid initial calibration following the procedures outlined in Section 10.3 or
          confirm that the calibration is still valid by running a CCC as described in Section 10.4.
          If establishing an initial calibration for the first time, complete the IDC as described in
          Section 9.2.

    11.3.3  Analyze aliquots of Field and QC Samples at appropriate frequencies (Section 9) with the
          LC/MS conditions used to acquire the initial calibration. At the conclusion of data
          acquisition, use the same software settings that were used in the calibration procedure to
          identify peaks in the predetermined  retention time windows.
   11.3.4 COMPOUND IDENTIFICATION - Establish an appropriate retention time window for
          the target analyte and internal standard so they may be identified in QC and field sample
          chromatograms. Because the retention time will vary with the ionic strength of the
          sample the absolute retention time windows may need to be wider than usual.

       11.3.4.1    High ionic strength samples will  cause retention times to decrease and introduce
                 band broadening.  This is unavoidable due to the high elutropic strength of the
                 matrix in these samples. Because of these effects, the relative retention time of
                 perchlorate is used as one  component for qualitative identification (Section 9.3.6).
                 The relative retention time must be 1.0 + 2% (0.98-1.02) for a peak to be
                 identified as perchlorate.

       11.3.4.2    Perchlorate has a strong M+2 ion due to the presence of 37C1. The analyst must
                 confirm that the ratio of the molecular ion to its M+2 ion (99/101) for the SIM
                 technique or the ratio of their daughter ions (83/85) for the MRM technique are
                 within ± 25% of the theoretical value of 3.08 (2.31 to 3.85). If this ion ratio
                 requirement is not met in any QC sample, then all samples in the Analysis Batch

                                         331.0-21

-------
                    are considered invalid and must be reanalyzed after reestablishing acceptable
                    instrument performance.

              11.3.4.2.1  If the isotope area ratio criteria are not met when using MRM detection the
                        analyst should report the associated values as "suspect due to lack of
                        confirmation."

              11.3.4.2.2  When using SIM detection, remedial action is required if the following
                        conditions exist. If a peak is detected at m/z 101, its concentration is > MRL,
                        the relative retention time criterion is met and the isotope area ratio
                        requirement is not met. In this case, sample pretreatment must be performed.
                        The most frequent cause of ratio failure is poor chromatographic resolution
                        between sulfate and perchlorate.  If the ratio failure is due to a high sulfate
                        level in the sample, dilution or sulfate removal using a commercial barium
                        pretreatment cartridge (Sect.  6.10) may be successful. Consult the
                        manufacturer's instructions for preparation of the barium pretreatment
                        cartridge prior to use with  samples. Generally, the procedure requires rinsing
                        the cartridge with a minimum volume of reagent water.  It has been found that
                        rinsing with approximately twice the recommended volume of water yields
                        better results. Add IS to the sample prior to sample pretreatment using the
                        cartridges.  Additionally, chromatographic conditions may be changed to
                        eliminate the coelution, however the IDC must be repeated using the new
                        conditions.

                        NOTE: If cartridges are used, a reagent water blank with IS should be
                        processed to insure that the cartridges don't contribute interferences or retain
                        perchlorate.

              11.3.4.2.3  Following pretreatment, reanalyze the sample as part of the same or a
                        subsequent Analysis Batch. If the ratio criteria of the reinjected sample pass,
                        report the results of the reinjected sample. If the ratio criteria are still not met
                        the analyst should report the  associated values as "suspect due to lack of
                        confirmation."

       11.3.5  EXCEEDING CALIBRATION RANGE -The analyst must not extrapolate beyond the
              established calibration range. If an analyte result exceeds the range of the initial
              calibration curve, the sample may be diluted with reagent water, with the appropriate
              amount of internal standard added to match the original level,  and the diluted sample
              reinjected. Incorporate the dilution factor into final  concentration calculations.  The
              dilution will  also affect analyte MRLs.

12. DATA ANALYSIS AND CALCULATIONS

   12.1  Identify the analyte in the Field and QC Samples as described in Section 11.3.4.

       12.1.1  Sulfate has an ion at m/z 99 (H34SO4) that can interfere with confirmation if the
              chromatography is not optimized or if the sulfate concentration in the native samples is
                                            331.0-22

-------
             very high. The SIM version of this method is significantly more susceptible to
             interferences from hydrogen sulfate than the MRM version of this method.  The M + 2
             ion (m/z 101) must be used for quantitation of perchlorate when using SIM.

   12.2  Calculate analyte concentrations using the multipoint calibration established in Section 10.3.
         Report only those values that fall between the MRL and the highest calibration standard.
         Samples with target analyte responses that exceed the highest standard require dilution and
         reanalysis (Sect. 11.3.5).

   12.3  Calculations must use all available digits of precision, but final reported concentrations should
         be rounded to an appropriate number of significant figures (one digit of uncertainty), typically
         two, and not more than three significant figures.

   12.4  Prior to reporting data, the laboratory is responsible for assuring that QC requirements have
         been met or that any appropriate qualifier is documented.

13. METHOD PERFORMANCE

   13.1  PRECISION, ACCURACY AND DETECTION LIMITS - Tables for these data are
         presented in  Section 17.  Instrumental conditions and monitored ions are presented in Table 1.
         The LCMRL for perchlorate is presented in Table 2 and was calculated using a procedure
         described elsewhere1. Single laboratory precision and accuracy data are presented in Tables 3
         and 4.  Table 5 summarizes the requirements for the initial demonstration of capability (IDC)
         and Table 6 summarizes the requirements for the required ongoing quality control. Figure 1
         shows a representative chromatogram from a MRM analysis. Figure 2 shows a representative
         chromatogram from a SIM analysis and Figure 3 shows a total ion chromatogram (m/z 99, 101
         and 107) of a LSSM that contains the common anions chloride, sulfate and bicarbonate at
         1000 mg/L.

   13.2  SECOND LABORATORY DEMONSTRATION -- The performance of this method was
         demonstrated by a second laboratory, with results similar to those reported in Section 17. The
         authors wish to acknowledge the work of the U.S. EPA National Risk Management Research
         Laboratory, Ada, OK and the U.S. EPA New England Laboratory, North Chelmsford, MA for
         their participation in the second laboratory demonstration.
14. POLLUTION PREVENTION

   14.1  For information about pollution prevention that may be applicable to laboratory operations,
         consult "Less is Better: Laboratory Chemical Management for Waste Reduction" available
         from the American Chemical Society's Department of Government Relations and Science
         Policy, 1155 16th Street N.W., Washington, D.C., 20036, or on-line at
         http://www.ups.edu/community/storeroom/Chemical Wastes/wastearticles.htm
                                           331.0-23

-------
15. WASTE MANAGEMENT

   15.1  The analytical procedures described in this method generate relatively small amounts of waste
         since only small amounts of reagents and solvents are used. The matrices of concern are
         finished drinking water or source water. However, the Agency requires that laboratory waste
         management practices be conducted consistent with all applicable rules and regulations, and
         that laboratories protect the air, water, and land by minimizing and controlling all releases
         from fume hoods and bench operations.  Also,  compliance is required with any sewage
         discharge permits and regulations, particularly  the hazardous waste identification rules and
         land disposal restrictions. For further information on waste management, see the publications
         of the American Chemical  Society's Laboratory Environment, Health & Safety Task Force  on
         the Internet at http://membership.acs.Org/c/ccs/labehs/publications.htm.  Or see "Laboratory
         Waste Minimization and Pollution Prevention," Copyright ©  1996 Battelle Seattle Research
         Center, which can be found at http://www.p2pays.org/ref/01/text/00779/index2.htm.
16. REFERENCES

1.  Statistical Protocol for the Determination of the Single-Laboratory Lowest Concentration Minimum
   Reporting Level (LCMRL) and Validation of the Minimum Reporting Level (MRL), available at
   www.epa.gov/OGWDW/methods/sourcalt.html

2.  Glaser, J.A., D.L. Foerst, G.D. McKee, S.A. Quave, and W.L. Budde, "Trace Analyses for
   Wastewaters", Environ.  Sci. Technol.. 15 (1981) 1426_1435.

3.  "Carcinogens - Working With Carcinogens," Department of Health, Education, and Welfare, Public
   Health Service, Center for Disease Control, National Institute for Occupational Safety and Health,
   Publication No. 77-206,  Aug. 1977.

4.  "OSHA Safety and Health Standards, General Industry," (29CFR1910), Occupational Safety and
   Health Administration, OSHA 2206, (Revised, January 1976).

5.  "Safety in Academic Chemistry Laboratories," American Chemical Society Publication, Committee
   on Chemical Safety, 3rd Edition, 1979.

6.  Blosse, P.T., Boulter, E.M., Sundaram, S., "Diminutive Bacteria Implications for Sterile Filtration",
   Pall Corporation, East Hills, NY.

7.  ASTM Annual Book of  Standards, Part II, Volume 11.01, D3370-82, "Standard Practice for
   Sampling Water," American Society for Testing and Materials, Philadelphia, PA, 1986.

8.  Xu, J., Y. Song, B. Min, L. Steinberg, and B.E. Logan. 2003. Microbial degradation of perchlorate:
   principles and applications. Environ. Engin. Sci, 20(5): 405-422.

9.  Lange's Handbook of Chemistry (13th Edition); Dean, J. A., Ed; McGraw-Hill: New York, NY, 1985.
                                           331.0-24

-------
17.   TABLES, DIAGRAMS, FLOWCHARTS, AND VALIDATION DATA
     TABLE 1.  INSTRUMENTAL CONDITIONS, RETENTION TIMES,
                QUANTITATION IONS

                   Micromass QuattroMicro Mass Spectrometer Settings:
Parameter
Ion Source
Polarity
Capillary Voltage
Cone Voltage
Extractor
RFLens
Source Temperature
Desolvation Temperature
Cone Gas Flow
Desolvation Gas Flow
LM 1 Resolution
HM 1 Resolution
Ion Energy
Entrance
Collision
Exit
LM 2 Resolution
HM 2 Resolution
Ion Energy
Multiplier
Gas Cell Pressure
Function dwell time
Data Smoothing (mean)
LC/MRM
Electrospray
Negative Ion
0.58 kV
40V
3V
0.3V
120 °C
320 °C
60 L/hr
N2 800 L/hr
15
15
0.6V
1
24
1
14
14
IV
650V
Ar 3.54e-3 mbar
300ms
Window/Smoothes, 3/2
LC/SIM
Electrospray
Negative Ion
0.58 kV
40V
3V
0.3V
120 °C
320 °C
60 L/hr
N2 800 L/hr
15
15
0.6V
50
2
50
15
15
IV
650V
Off
300ms
Window/Smoothes, 3/2
      Chromatographic conditions: LC; Waters 2690, column; Dionex IonPak®AS-21 2.1 x 150 mm
      7 |j,m particle diameter, injection volume; 100 |jL, mobile phase; 200 mM methylamine isocratic
      at 350 |jL/min.
                             Monitored Ions and Retention Times
Analyte
J5C103
37C103
35C118O3 (IS)
*Precursor Ion
(m/z)
99
101
107
fProduct Ion
(m/z)
83
85
89
RT
(min)
8.68
8.68
8.65
            *In SIM analysis only the precursor ions are monitored and m/z 101 must be used for quantitation.
            In MRM analysis either perchlorate product ion may be used for quantitation.
                                          331.0-25

-------
TABLE 2. LOWEST CONCENTRATION MRL AND DLs FOR PERCHLORATE
Analysis Method
LC/MRM
LC/SIM
Analyte
C104
CIO"
LCMRL (ng/L)
0.022
0.056
*DL (ng/L)
a0.005
b0.008
 *The DL was calculated from data acquired on a single day using m/z 83 (MRM) or m/z 101 (SIM)
 "Fortification 0.010
 fortification 0.075
TABLE 3. LC/MRM PRECISION AND RECOVERY DATA FOR PERCHLORATE IN
          VARIOUS MATRICES (n=7 REPLICATES)
Matrix
Reagent Water


Surface Water A

Surface Water B

Ground Water

* Synthetic High Ionic Strength

Unfortified
Concentration
(Hg/L)

-------
TABLE 5.  INITIAL DEMONSTRATION OF CAPABILITY QUALITY CONTROL
           REQUIREMENTS
  Method
 Reference
   Requirement
    Specification and
       Frequency
    Acceptance Criteria
 Section
 9.2.1
Demonstration of
Low System
Background
Analyze a LRB and
LSSMB prior to any other
IDC steps.
Demonstrate that the target
analyte is < 1/3 of the MRL
and that possible interferences
from sampling devices do not
prevent the identification and
quantitation of perchlorate.
 Section
 9.2.1.1
Test For System
Carryover
Analyze a reagent water
blank after the high CAL
standard during the IDC
calibration.
Demonstrate that the target
analyte is < 1/3 of the MRL
and that carry-over from
previous samples does not
prevent the identification and
quantitation of perchl orate.
 Section
 9.2.2
Demonstration of
Precision
Analyze 7 replicate LFBs
and 7 replicate LFSSMs
fortified near the midrange
concentration.
%RSD must be < 20%
 Section
 9.2.3
Demonstration of
Accuracy
Calculate average recovery
for replicates used in
Section 9.2.2.
Mean recovery 80-120% of the
true value
 Section
 9.2.4
Minimum
Reporting Limit
(MRL)
Confirmation
Fortify and analyze 7
replicate LFBs at the
proposed MRL
concentration. Calculate
the mean and the Half
Range (HR). Confirm that
the Upper PIR and Lower
PIR (Sect.  9.2.4.2) meet
the recovery criteria.	
                                                           Upper PIR < 150%

                                                           Lower PIR > 50%
                                           331.0-27

-------
TABLE 6.   ONGOING QUALITY CONTROL REQUIREMENTS (SUMMARY)
  Method
 Reference
   Requirement
     Specification and
        Frequency
     Acceptance Criteria
 Section
 8.3
Sample Holding
Time
28 days when processed and
stored according to Sections
8.1 and 8.2.
Sample results are valid only if
samples are analyzed within the
sample holding time.
 Section
 10.3
Initial Calibration
Use the internal standard
calibration technique to
generate a linear or quadratic
calibration curve.  Use at
least 5 standard
concentrations.  Check the
calibration curve as
described in Section 10.3.3.
When each calibration standard
is calculated as an unknown
using the calibration curve, the
lowest level standard should be
within 50-150% of the true
value. All other points should
be within 80-120% of the true
value.
 Section
 9.3.1
Laboratory Reagent
Blank (LRB)
Daily, or with each Analysis
Batch of up to 20 field
samples, whichever is more
frequent.
Results must be < 1/3 the MRL.
If the background exceeds 1/3
the MRL, the results for
perchlorate in the Analysis
Batch are invalid.
 Section
 10.4
Continuing
Calibration Check
(CCC)
Verify initial calibration by
analyzing a Low CCC and a
Mid CCC at the beginning of
each Analysis Batch.
Subsequent CCCs are
required after every 10 field
samples,  and after the last
field sample in a batch.
The lowest level CCC must be
within 50-150% of the true
value. All other points must be
within 80-120% of the true
value.

Results for field samples that
are not bracketed by acceptable
CCCs are invalid.
                                           331.0-28

-------
TABLE 6. (CONTINUED)
  Method
 Reference
   Requirement
     Specification and
        Frequency
     Acceptance Criteria
 Section
 9.3.3
Laboratory
Fortified Blank
(LFB)
Daily, or with each Analysis
Batch of up to 20 field
samples, whichever is more
frequent.
For LFBs fortified at
concentrations < MRL, the
result must be within 50-150%
of the true value. All other
LFBs must be within 80-120%
of the true value.
 Section
 9.3.4
Internal Standard
(IS)
Compare IS area to the IS
area of the first CCC in the
Analysis Batch for all QC
and field samples.
Peak area counts for the IS must
be within + 30% of the IS area
of the first CCC of the Analysis
Batch. If the IS area does not
meet this criterion, the
corresponding perchlorate
results are invalid.
 Section
 9.3.5
Isotope Area Count
Ratio
Monitor the isotope ratio for
all QC and field samples in
the Analysis Batch.
The calculated area count ratio
(m/z 99/101 for SIM, or m/z
85/83 MRM) must be within
±25% (2.31-3.85) of the
theoretical value of 3.086.
 Section
 9.3.6
Relative Retention
Time
Monitor the relative retention
time for all QC and field
samples in the Analysis
Batch.
The relative retention time for
perchlorate vs. the IS must be
between 0.98-1.02.
 Section
 9.3.7
Laboratory
Fortified Sample
Matrix (LFSM)
Analyze one LFSM per
Analysis Batch. Fortify the
LFSM with perchlorate at a
concentration close to but
greater than the native
concentration (if known).
Calculate LFSM recoveries.
For LFSMs fortified at
concentrations < MRL, the
result must be within 50-150%
of the true value. All other
LFSMs must be within 80-
120% of the true value.
                                           331.0-29

-------
TABLE 6.  (CONTINUED)
  Method
 Reference
   Requirement
     Specification and
        Frequency
     Acceptance Criteria
 Section
 9.3.8
Laboratory
Fortified Sample
Matrix Duplicate
(LFSMD) or
Laboratory
Duplicate(LD)
Analyze at least one LFSMD
or LD daily, or with each
Analysis Batch of up to 20
field samples, whichever is
more frequent.
For LFSMDs or LDs having
perchlorate concentrations < 2 x
MRL, the RPD must be within
50-150%. All other LFSMDs
and LDs must have RPDs
within 80-120%.
 Section
 10.3.4
Quality Control
Sample (QCS)
Analyzed with every new
calibration curve.
Results must be 80-120% of the
true value.
                                         331.0-30

-------
TABLE 7. EXAMPLE SEQUENCE FOR AN ANALYTICAL BATCH
Sample
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
Sample Description
CCC < MRL
Mid Level CCC
LRB
LFB
Field Sample- 1
Field Sample-2
Field Sample-3
Field Sample-4
Field Sample-5
Field Sample-6
Field Sample-7
Field Sample-8
Field Sample-9
Field Sample- 10
CCC (rotate level)
LFSM
LD or LFSMD
Field Sample- 11
Field Sample-12
Field Sample- 13
Field Sample-14
Field Sample- 15
Field Sample- 16
Field Sample- 17
Field Sample- 18
Field Sample- 19
Field Sample-20
CCC (rotate level)
*Acceptance Criteria
5 0-1 50% of true value
80- 120% true value
< 1/3 MRL
< MRL 50-150% of true value
> MRL 80-120% of true value










< MRL 50-150% of true value
> MRL 80-120% of true value
< MRL 50-150% of true value
> MRL 80-120% of true value
<2xMRL50-150%RPD
> 2 x MRL 80-120% RPD










< MRL 50-150% of true value
> MRL 80-120% of true value
              * Additionally, all samples must meet the isotope area count ratio
              criteria (Section 9.3.5) and the relative retention time criteria (Section
              9.3.6).
                                            331.0-31

-------
                                                                 FIGURE 1
                             LC/MRM CHROMATOGRAM OF A 0.5 ^g/L PERCHLORATE STANDARD
04-25-04-69
 100-,
                                                       Internal Standard m/z 107>89
MRM of 3 Channels ES-
            107 > 89
             6.74e3
04-25-04-69
 100-1
                                                      Perchlorate  m/z 101>85
MRM of 3 Channels ES-
            101 >85
             1.58e3
04-25-04-69
 100-,
                                                      Perchlorate  m/z 99>83
MRM of 3 Channels ES-
             99>83
             4.10e3
                                                                                                                                     	Time
  0.00
              1.00
                         2.00
                                    3.00
                                               4.00
                                                          5.00       33ao&-32      7.00
                                                                                            8.00
                                                                                                       9.00
                                                                                                                  10.00
                                                                                                                             11.00
                                                                                                                                        12.00

-------
                                                                     FIGURE 2
                                LC/SIM CHROMATOGRAM OF A 0.5 ^g/L PERCHLORATE STANDARD
07-13-04-07
 100-1
  0-^^
                                                        Internal Standard m/z 107
                                                               SIR of 3 Channels ES-
                                                                              107
                                                                            1.13e5
07-13-04-07
 100-1
                                                        Perchlorate m/z 101
                                                               SIR of 3 Channels ES-
                                                                              101
                                                                            3.78e4
07-13-04-07
 100-,
                                                        Perchlorate m/z 99
                                                               SIR of 3 Channels ES-
                                                                               99
                                                                            1.01e5
                                                                                                                                               Time
            4.00       4.50        5.00        5.50
                                                        6.00
6.50        7.00
     331.0-33
                                                                                         7.50
                                                                                                    8.00       8.50       9.00       9.50

-------
07-13-04-01

 100n
                                                               FIGURE 3
                          LC/SIM TIC CHROMATOGRAM (m/z  99,101,107) OF A LFSSM (1000 mg/L)
                                           FORTIFIED WITH 0.5 (ig/L PERCHLORATE
SIR of 3 Channels ES-
              TIC
            9.89e5
                                                                 331.0-34
                                                                                                  /     \
         0.50    1.00    1.50    2.00   2.50    3.00    3.50   4.00    4.50    5.00   5.50    6.00    6.50   7.00    7.50   8.00    8.50    9.00   9.50
                                                                                                                                         Time

-------