METHOD 552.3    DETERMINATION OF HALOACETIC ACIDS AND DALAPON IN
                  DRINKING WATER BY LIQUID-LIQUID MICROEXTRACTION,
                  DERIVATIZATION, AND GAS CHROMATOGRAPHY WITH
                  ELECTRON CAPTURE DETECTION
                                  EPA 815-B-03-002


                                     Revision 1.0

                                      July 2003
M. M. Domino and B.V. Pepich (Shaw Environmental and Infrastructure, Inc.)
DJ. Munch and P.S. Fair (US EPA, Office of Ground Water and Drinking Water)
Y. Xie (Penn State University)

D. J. Munch, J.W. Munch (US EPA, Office of Ground Water and Drinking Water) and A. M. Pawlecki-
Vonderheide (ICI) Method 552.2, Revision 1.0 (1995)

J.W. Hodgeson (USEPA), D. Becker (Technology Applications, Inc.) Method 552.1, (1992)

J.W. Hodgeson (USEPA), J. Collins and R. E. Earth (Technology Applications, Inc.) Method 552.0,
(1990)
                           TECHNICAL SUPPORT CENTER
                OFFICE OF GROUND WATER AND DRINKING WATER
                   U. S. ENVIRONMENTAL PROTECTION AGENCY
                              CINCINNATI, OHIO 45268
                                       552.3-1

-------
                                METHOD 552.3

 DETERMINATION OF HALO ACETIC ACIDS AND DALAPON IN DRINKING WATER BY
          LIQUID-LIQUID MICROEXTRACTION, DERIVATIZATION, AND GAS
            CHROMATOGRAPHY WITH ELECTRON CAPTURE DETECTION
1.     SCOPE AND APPLICATION

      1.1    This is a gas chromatography (GC) method for the determination of haloacetic acids and
             dalapon in drinking waters. Accuracy, precision, and Detection Limit (DL) data are
             presented for the following compounds in reagent water and finished ground and surface
             waters.
                         Analyte

           Bromochloroacetic acid (BCAA)
           Bromodichloroacetic acid (BDCAA)
           Chlorodibromoacetic acid (CDBAA)
           Dalapon
           Dibromoacetic acid (DBAA)
           Dichloroacetic acid (DCAA)
           Monobromoacetic acid (MBAA)
           Monochloroacetic acid (MCAA)
           Tribromoacetic acid (TBAA)
           Trichloroacetic acid (TCAA)
Chemical Abstracts Service
          (CAS)
     Registry Number
        5589-96-8
        71133-14-7
        5278-95-5
         75-99-0
        631-64-1
         79-43-6
         79-08-3
         79-11-8
         75-96-7
         76-03-9
      1.2    Detection Limits are compound, instrument, and matrix dependent. The DL is defined as
             the statistically calculated minimum amount that can be measured with 99% confidence
             that the reported value is greater than zero.(1) Detection Limits for the above listed
             analytes are provided in Section 17, Table 5.  The DL differs from, and is lower than, the
             Minimum Reporting Level (MRL) (Sect. 3.17). The concentration range for target
                                         552.3-2

-------
             analytes in this method was evaluated between 0.5 jig/L and 30 jig/L for a 40-mL sample.
             Precision and accuracy data are presented in Section 17, Tables 6-11.

       1.3    This method is restricted to use by or under the supervision of analysts skilled in liquid-
             liquid extractions, derivatization procedures and the use of GC and interpretation of gas
             chromatograms.
2.     SUMMARY OF METHOD

2.1     A 40-mL volume of sample is adjusted to a pH of 0.5 or less and extracted with 4 mL of either
       methyl tert-butyl ether (MTBE) or tert-amyl methyl ether (TAME) containing an internal
       standard. The haloacetic acids that have been partitioned into the organic phase are then
       converted to their methyl esters by the addition of acidic methanol followed by heating for 2
       hours. The solvent phase containing the methylated haloacetic acids is separated from the acidic
       methanol by adding 7 mL of a concentrated aqueous solution of sodium sulfate. The aqueous
       phase is discarded.  The extract is then neutralized with a saturated solution of sodium
       bicarbonate and the solvent layer is removed for analysis. The target analytes are identified and
       quantified by capillary column gas chromatography using an electron capture detector
       (GC/ECD). Analytes are quantified using procedural standard calibration.
3.     DEFINITIONS

       3.1    EXTRACTION BATCH - A set of up to 20 Field Samples (not including calibration
             standards and QC samples) extracted together by the same person(s) during a work day
             using the same lots of solvents, surrogate solution, and fortifying solutions.  Required QC
             samples include a Laboratory Reagent Blank, Laboratory Fortified Matrix, either a Field
             Duplicate or Laboratory Fortified Matrix Duplicate and an appropriate number of
             Continuing Calibration Checks.

       3.2    ANALYSIS BATCH - A set of samples that is analyzed on the same instrument during a
             24-hour period that begins and ends with the analysis of the appropriate Continuing
             Calibration Check standards  (CCC). Additional CCCs may be required depending on the
             length of the analysis batch and/or the number of Field Samples.
                                            552.3-3

-------
3.3    INTERNAL STANDARD (IS) - A pure analyte added to an extract or standard solution
      in a known amount and used to measure the relative responses of other method analytes
      and surrogates.  The internal standard must be an analyte that is not a sample component.

3.4    SURROGATE ANALYTE (SUR) - A pure analyte, which chemically resembles target
      analytes and is extremely unlikely to be found in any sample.  This analyte is added to a
      sample aliquot in a known amount before extraction or other processing, and is measured
      with the same procedures used to measure other sample components.  The purpose of the
      SUR is to monitor method performance with each sample.

3.5    LABORATORY REAGENT BLANK (LRB) - An aliquot of reagent water or other
      blank matrix that is treated exactly as a sample including exposure to all glassware,
      equipment, solvents, reagents, sample preservatives, internal standards, and surrogates
      that are used in the extraction batch. The LRB is used to determine if method analytes or
      other interferences are present in the laboratory environment, the reagents, or the
      apparatus.

3.6    LABORATORY FORTIFIED BLANK (LFB) - An aliquot of reagent water or other
      blank matrix to which known quantities of the method analytes and all the preservation
      compounds are added. The LFB is analyzed exactly like a sample, and its purpose is to
      determine whether the methodology is in control, and whether the laboratory is capable of
      making accurate and precise measurements.

3.7    LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - An aliquot of a Field
      Sample to which known quantities of the method analytes and all the preservation
      compounds are added. The LFSM is processed and analyzed exactly like a sample, and
      its purpose is to determine whether the sample matrix contributes bias to the analytical
      results. The background concentrations of the analytes in the sample matrix must be
      determined in a separate aliquot and the measured values in the LFSM corrected for
      background concentrations.

3.8    LABORATORY FORTIFIED SAMPLE MATRIX DUPLICATE  (LFSMD) - A second
      aliquot of the Field Sample used to prepare the LFSM which is fortified, extracted and
      analyzed identically to the LFSM. The LFSMD is used instead of the Field Duplicate to
      access method precision and accuracy when the occurrence of a target analyte is
      infrequent.
                                    552.3-4

-------
3.9    LABORATORY DUPLICATES (LD1 and LD2) - Two aliquots of the same sample split
       in the laboratory and analyzed separately with identical procedures. Analyses of LD1 and
       LD2 indicate precision associated with laboratory procedures, but not with sample
       collection, preservation, and storage procedures.

3.10   FIELD DUPLICATES (FD1 and FD2) - Two separate samples collected at the same
       time and place under identical circumstances, and treated exactly the same throughout
       field and laboratory procedures. Analyses of FD1 and FD2 give a measure of the
       precision associated with sample collection, preservation, and storage, as well as with
       laboratory procedures.

3.11   STOCK STANDARD SOLUTION (SSS) - A concentrated solution containing one or
       more method analytes prepared in the laboratory using assayed reference materials or
       purchased from a reputable commercial source.

3.12   PRIMARY DILUTION STANDARD (PDS) SOLUTION - A solution containing
       method analytes prepared in the laboratory from stock standard solutions and diluted as
       needed to prepare calibration solutions and other  needed analyte solutions.

3.13   CALIBRATION STANDARD (CAL) - A solution prepared from the primary dilution
       standard solution and/or stock standard solutions  and the internal standards and surrogate
       analytes.  The CAL solutions are used to calibrate the instrument response with respect to
       analyte concentration.

3.14   CONTINUING CALIBRATION CHECK (CCC) - A calibration standard containing the
       method analytes, internal standard(s) and surrogate(s), which is analyzed periodically to
       verify the accuracy of the existing calibration for  those analytes.

3.15   QUALITY CONTROL SAMPLE (QCS) -A sample prepared using a PDS of method
       analytes that is obtained from a source external to the laboratory and different from the
       source of calibration standards.  The second source PDS and the surrogate PDS are used
       to fortify the QCS at a known concentration. The QCS is used to check calibration
       standard integrity.

3.16   DETECTION LIMIT - The minimum concentration of an analyte that can be identified,
       measured and reported with 99% confidence that  the analyte concentration is greater than
       zero (Sect. 9.2.4). This is a statistical determination of precision. Accurate quantitation
       is not expected at this level.(1)

                                    552.3-5

-------
       3.17   MINIMUM REPORTING LEVEL (MRL) - The minimum concentration that can be
             reported as a quantified value for a target analyte in a sample following analysis.  This
             defined concentration can be no lower than the concentration of the lowest continuing
             calibration standard for that analyte, and can only be used if acceptable quality control
             criteria for this standard are met.

       3.18   PROCEDURAL STANDARD CALIBRATION - A calibration method in which
             aqueous calibration standards are prepared and processed (e.g., extracted, and/or
             derivatized) in exactly the same manner as the samples. All steps in the process from
             addition of sampling preservatives through instrumental analyses are included in the
             calibration. Using procedural standard calibration compensates for any inefficiencies in
             the processing procedure.

       3.19   MATERIAL SAFETY DATA SHEET (MSDS) - Written information provided by
             vendors concerning a chemical's toxicity, health hazards, physical properties, fire, and
             reactivity  data including storage, spill, and handling precautions.
4.      INTERFERENCES

       4.1    All glassware must be meticulously cleaned. Wash glassware with detergent and tap
             water, rinse with tap water followed by reagent water.  A final rinse with solvents may be
             needed. In place of a solvent rinse, non-volumetric glassware can be muffled at 400 °C
             for 2 hours. Volumetric glassware should not be heated in an oven above 120 °C.  Store
             glassware inverted or capped with aluminum foil.

       4.2    Method interferences may be caused by contaminants in solvents, reagents (including
             reagent water), sample bottles and caps, and other sample processing hardware that lead
             to discrete artifacts and/or elevated baselines in the chromatograms. All items used in the
             method must be routinely demonstrated to be free from interferences (less than V3 the
             MRL for each target) under method conditions by processing and analyzing laboratory
             reagent blanks as described in Section 9.  Subtracting blank values from sample
             results is  not permitted.

             4.2.1   Sodium sulfate can be a source of method interferences. After screening several
                    brands, it was found that a grade suitable for pesticide residue analysis had the
                    lowest amount of method interferants. If the suitability of the available sodium

                                            552.3-6

-------
              sulfate is in question, extract and analyze a laboratory reagent blank (Sect. 3.5) to
              test for interferences prior to sample processing.

       4.2.2   The ester of bromochloroacetic acid (BCAA) coelutes with a low-level interferant
              on both the primary and confirmation column. This interferant is present in the
              LRB and believed to be dimethyl sulfide that originates in the sodium sulfate.
              Because of difference in peak shape, the BCAA ester tends to "ride on" the
              interferant allowing accurate quantitation. Concentrations of BCAA at or below
              2|ig/L may require manual integration of the BCAA.

       4.2.3   TAME, which has higher methylation efficiencies for the trihaloacetic acids than
              MTBE, is  currently only available at 97% purity. Different lots of TAME should
              be carefully evaluated for potential interferants.  Several lots of TAME contained
              an interferant that coeluted with the ester of monochloroacetic acid (MCAA) on
              the primary column.

4.3    The esters of the surrogate used in Method  552.2 (2,3-dibromopropanoic acid) and
       dichlorobromoacetic acid are not completely resolved under the chromatographic
       conditions for the  confirmation column listed in Tables 3 and 4. The current surrogate,
       2-bromobutanoic acid, is fully resolved from all method compounds on both columns
       under the conditions listed in Tables 1-4 (Figs. 1-4).

4.4    Matrix interferences may be  caused by contaminants that are extracted from the sample.
       The extent of matrix interferences will vary considerably from source to source,
       depending upon the water sampled.

4.5    Interferences by phthalate esters can pose a major problem in analysis when using an
       electron capture detector (ECD). These compounds generally appear in the chromatogram
       as large peaks. Common flexible plastics contain varying amounts of phthalates that are
       easily extracted or leached during laboratory operations. Cross contamination of clean
       glassware routinely occurs when plastics are handled during extraction steps, especially
       when solvent-wetted surfaces are handled. Interferences from phthalates can best be
       minimized by avoiding the use of plastics in the laboratory. Exhaustive purification of
       reagents and glassware may be required to eliminate background phthalate
       contamination.(2>3)
                                     552.3-7

-------
5.      SAFETY
       5.1    The toxicity or carcinogenicity of each reagent used in this method has not been precisely
             defined. Each chemical compound should be treated as a potential health hazard, and
             exposure to these chemicals should be minimized.  The laboratory is responsible for
             maintaining a current awareness file of OSHA regulations regarding the safe handling of
             the chemicals specified in this method.  A reference file of MSDSs should also be made
             available to all personnel involved in the chemical analysis.  Additional references to
             laboratory safety are available.(4~6)

       5.2    Pure standard materials and stock standards of these compounds should be handled with
             suitable protection to skin and eyes, and care should be taken not to breathe the vapors or
             ingest the materials.

       5.3    The toxi cities of the extraction solvents, MTBE or TAME, have not been well defined.
             Susceptible individuals may experience adverse effects upon skin contact or inhalation of
             vapors. Therefore protective clothing and gloves should be used and MTBE or TAME
             should be used only in a chemical fume hood or glove box. The same precaution applies
             to pure standard materials.
6.     EQUIPMENT AND SUPPLIES

      6.1    SAMPLE CONTAINERS - Amber glass bottles, at least 50-mL, fitted with PTFE
             (polytetrafluoroethylene) lined screw caps.

      6.2    EXTRACTION VIALS - 60-mL clear glass vials with PTFE-lined screw caps.

      6.3    AUTOSAMPLER VIALS - 2.0-mL amber vials with screw or crimp cap and a PTFE-
             faced seal.

      6.4    STANDARD SOLUTION STORAGE CONTAINERS - 10- to 20-mL amber glass vials
             with PTFE-lined screw caps.

      6.5    GRADUATED CONICAL CENTRIFUGE TUBES WITH PTFE-LINED SCREW
             CAPS - 15-mL with graduated 1-mL markings (Fisher Cat. #: 05-538-30A or
             equivalent).
                                          552.3-8

-------
6.6   PASTEUR PIPETTES - Glass, disposable.

6.7   PIPETTES - Class A, 2.0-, 3.0-, 4.0-, and 7.0-mL glass, or adjustable volume dispensers.

6.8   VOLUMETRIC FLASKS - Class A, suggested sizes include 5-mL, 10-mL, 100-mL.

6.9   MICRO SYRINGES - Various sizes.

6.10  HEATER (BLOCK,or SAND BATH or WATER BATH) - A block heater (Thermolyne
      Model DB16525, Fisher Cat. #: 11-716-50 or equivalent) capable of maintaining
      regulated, elevated temperature equipped with a sand bath or heating block (Fisher Cat. #:
      11-716-27 or equivalent) capable of holding screw cap conical centrifuge tubes in Section
      6.5 was used for method development. A thermostated water bath equipped with a
      surface layer of small plastic spheres (Fisher Cat. #.  14-220-31 or equivalent) to retard
      evaporation and subsequent heating of tube walls may  also be used.

6.11  pH PAPER - With a pH range of 0 - 1.5 (Fisher Cat. #. 14-853-55 or equivalent).

6.12  BALANCE - Analytical, capable of weighing to the nearest 0.0001 g.

6.13  VORTEXER - Used to mix sample extracts (VWR Vortex-Genie, Cat. #: 14216-184 or
      equivalent).

6.14  GAS CHROMATOGRAPH - Capillary GC, with a low volume (150 |iL) micro ECD
      (Agilent Model 6890 or equivalent). The injector system should not allow analytes to
      contact hot stainless steel or other metal surfaces that promote decomposition. The
      performance data in Section 17  was collected using hot, splitless injection using a 2-mm
      i.d. quartz liner. Other injection techniques such as temperature programmed injection,
      cold on-column injection, and large volume injection may be used if the QC criteria of
      Sections 9 and 10 are met.  Alternate detectors which have  equivalent or greater
      selectivity for the target compounds may be  used.

6.15  AUTOSAMPLER - Agilent Model 7683 or equivalent.

6.16  PRIMARY GC COLUMN - DB-1701, 30-meter length, 0.25-mm i.d., 0.25-|im film,
      fused silica capillary  with chemically bonded (14%
      cyanopropylphenyl-methylpolysiloxane), or equivalent bonded, fused silica column.
                                    552.3-9

-------
       6.17   CONFIRMATION GC COLUMN-DB-5.625, 30-meter length, 0.25-mm i.d., 0.25-|im
             film, fused silica capillary with chemically bonded ("equivalent to" 5% phenyl-
             methylpolysiloxane), or equivalent bonded, fused silica column.
7.      REAGENTS AND STANDARDS

       7.1    REAGENTS AND SOLVENTS - Reagent grade or better chemicals should be used in
             all analyses. Unless otherwise indicated, it is intended that all reagents shall conform to
             the specifications of the Committee on Analytical Reagents of the American Chemical
             Society, where such specifications are available. Other grades may be used, provided it is
             first determined that the reagent is of sufficiently high purity to permit its use without
             lessening the quality of the determination.

             7.1.1   REAGENT WATER - Purified water which does not contain any measurable
                    quantities of any target analytes or interfering compounds greater than 1/3 the
                    MRL for each compound of interest.

             7.1.2   METHYL tert-BUTYL ETHER (MTBE) - High purity, demonstrated to be free
                    from analytes and interferences (HPLC grade or better).

             7.1.3   Tert-AMYL METHYL ETHER (TAME) - High purity, demonstrated to be free
                    from analytes and interferences.

                    7.1.3.1 Currently, TAME is available in > 97% purity.  Several manufacturers'
                          products were evaluated and found to contain a compound that elutes with
                          MCAA on the primary column. Lots evaluated from Fluka (Riedel-de
                          Haen), stored over molecular sieves, were free from this interferant.

             7.1.4   SODIUM SULFATE, Na2SO4 - Pesticide grade, granular, anhydrous.
                    Interferences have been observed when lower quality grades have been used. If
                    interferences are observed, it may be necessary to heat the sodium sulfate in a
                    shallow tray at 400 °C for up to 4 hr to remove phthalates and other interfering
                    organic substances. Store in a capped glass bottle rather than a plastic container.

             7.1.5   SODIUM BICARBONATE, NaHCO3 - ACS reagent grade.

             7.1.6   AMMONIUM CHLORIDE, NB^Cl - ACS reagent grade.

                                          552.3-10

-------
      7.1.7   SULFURIC ACID - Concentrated, ACS reagent grade. Substitution of
             hydrochloric acid (HC1) is not allowed. Solutions of HC1 can contain trace
             levels of bromide, which can promote the formation of brominated HAAs.

      7.1.8   HELIUM (or HYDROGEN)- 99.999% pure or better, GC carrier gas.

      7.1.9   ARGON 95%/METHANE 5% (or NITROGEN) - 99.999% pure or better, ECD
             make-up gas.

      7.1.10 SODIUM SULFATE SOLUTION - Prepare an  aqueous solution of sodium
             sulfate in reagent water at a concentration of 150 g/L.  Substitution of sodium
             chloride is not allowed. Sodium chloride can  contain trace levels of bromide,
             which can promote the formation of brominated HAAs.

      7.1.11 SODIUM BICARBONATE SOLUTION, SATURATED - Add sodium
             bicarbonate to a volume of water, mixing periodically until the solution has a
             small amount of undissolved sodium bicarbonate that does not disappear upon
             further mixing.

      7.1.12 10% SULFURIC ACID INMETHANOL SOLUTION-Add lOmL of sulfuric
             acid dropwise  (due to heat evolution) to 50-60 mL of methanol contained in a
             100-mL volumetric flask that has been placed in a cooling bath.  Mix, let cool,
             and dilute to the  100-mL mark with methanol.

7.2    STANDARD SOLUTIONS - Standard Solutions may be prepared from certified,
      commercially available solutions or from neat compounds. When preparing from neat
      material, compound purity needs to be 96% or greater.  When a compound purity is
      assayed to be 96% or greater, the weight can be used without correction to calculate the
      concentration of the stock standard. Solution concentrations listed in this section were
      used to develop this method and are included as an example.  Standards for sample
      fortification generally should be prepared in the smallest volume that can be accurately
      measured to minimize the addition of organic  solvent to aqueous samples. Laboratories
      should use standard QC  procedures to determine when Standard Solutions described in
      this section need to be replaced.

      7.2.1   STANDARD PREPARATION AND STORAGE TECHNIQUES - All Standard
             Solutions are prepared from either neat or solid materials following the general
             procedure outlined below.  This procedure assumes that the standard stock

                                   552.3-11

-------
       solutions are prepared in 10-mL volumetric flasks.  The procedure should be
       scaled accordingly if larger volumetric flasks are used. Most standards in this
       method are made up in MTBE, whether the extracting solvent is MTBE or
       TAME.  This is because MTBE is more water-soluble than TAME.

       7.2.1.1 For analytes which are solids in their pure form, prepare stock standard
             solutions by accurately weighing the suggested amount (refer to the
             sections below for suggested weights) of pure material into a clean, tared
             10-mL volumetric flask. Dilute the flask to volume with MTBE and mix
             thoroughly.

       7.2.1.2 For analytes which are liquid in their pure form at room temperature, place
             about 9.8 mL of MTBE into a 10-mL volumetric flask. Allow the flask to
             stand, unstoppered, for about 10 minutes to allow solvent film to evaporate
             from the inner walls of the volumetric, and weigh to the nearest 0.1 mg.
             Using a 10-jiL (or gaslight) syringe add the desired volume of the neat
             standard material to the flask by keeping the syringe needle just above the
             surface of the MTBE.  Be sure that the standard material falls dropwise
             directly into the MTBE without contacting the inner wall of the
             volumetric.  Record the weight and calculate the concentration of the stock
             standard from the net gain in weight. Dilute to volume, stopper, then mix
             by inverting the flask several times.

       7.2.1.3 Transfer the stock solutions to amber glass vials or amber bottles with
             PTFE-lined caps and store at < 0 °C.

7.2.2   INTERNAL STANDARD (IS) SOLUTIONS - 1,2,3-trichloropropane (99+%) is
       used as an internal standard for the method.  This compound has been shown to be
       an effective internal standard for the method analytes, but other compounds may
       be used if the QC requirements in Section 9 are met.

       7.2.2.1 INTERNAL STANDARD STOCK SOLUTION (2.0 mg/mL) - Prepare
             an internal standard stock solution by accurately transferring
             approximately 0.0200 g of neat 1,2,3-trichloropropane (weighed to the
             nearest O.lmg) into a 10-mL volumetric flask  containing methanol  as
             described in Section 7.2.1. The resulting concentration of the stock
             internal standard solution will be approximately 2.0 mg/mL.
                             552.3-12

-------
       7.2.2.2 INTERNAL STANDARD PRIMARY DILUTION STANDARD
             (100|ig/mL) - Prepare an internal standard fortification solution at
             100|ig/mL (or other suitable concentration) by the addition of 500 jiL (or
             other appropriate volume) of the stock standard to a 10-mL volumetric
             flask containing MTBE. Dilute to volume, mix thoroughly and transfer to
             an amber glass vial with a PTFE-lined screw cap and store at < 0°C.

       7.2.2.3 MTBE (or TAME) EXTRACTION SOLVENT WITH INTERNAL
             STANDARD (1.00 |ig/mL) - The internal standard 1,2,3-trichloropropane
             is added to the extraction solvent prior to analyte extraction to compensate
             for any volumetric differences encountered during sample processing.
             Have enough working solvent available to extract all calibration standards,
             QC samples and field samples in each extraction batch. The volume of
             fortified solvent should be determined by the sample workload. Never use
             two different lots of working solvent for one extraction batch. The
             following example illustrates preparation of 100 mL of fortified solvent.
             Add 1 mL of the primary dilution standard (100 |ig/mL) to a  100-mL
             volumetric flask containing MTBE (or TAME), dilute to volume and mix
             thoroughly.  Transfer the standard to an amber bottle for storage.  This
             results in a final internal standard concentration of 1.00 |ig/mL. This
             solution is used to extract the samples (Sect. 11.1).

7.2.3   SURROGATE (SUR) ANALYTE STANDARD SOLUTION - 2-bromobutanoic
       acid (99+%) is used as a surrogate compound in this method to evaluate the
       extraction and derivatization procedures.  This compound has been shown to be
       an effective surrogate for the method analytes and is well resolved from all target
       analytes and common interferants on both chromatographic columns. The Method
       552.2 surrogate, 2,3-dibromopropanoic acid,  can be used with this method, but
       requires a longer chromatographic run to separate it from DCBAA (Sect.  4.3).
       Other SUR analytes may be used as long as they are halogenated carboxylic acids
       and the QC requirements of Section 9.7 are met. All surrogates must be added as
       free acids. Alternate candidates should be confirmed to have adequate
       esterification efficiencies (e.g. > 80%) by comparison to a commercially available
       premethylated standard.  Lower esterification efficiencies may lead to poor SUR
       precision.

       Note: 2-bromo-2-methylpropanoic acid was ruled out as a potential SUR because
       it was found to degrade in the Field and QC Samples if they were not processed

                             552.3-13

-------
       immediately. Nearly 50% of the fortified concentration was lost when samples
       were fortified and then allowed to set for 3 hours prior to acidification. Alternate
       surrogate compounds MUST be carefully evaluated.

       7.2.3.1 SURROGATE STOCK SOLUTION (10 mg/mL) - Prepare surrogate
             stock standard solutions of 2-bromobutanoic acid [80-58-0] by accurately
             transferring approximately 0.100 g of the neat material (weighed to the
             nearest 0.1 mg.) into a clean, tared  10-mL volumetric flask as described in
             Section 7.2.1.  The resulting concentration of the Surrogate Stock Solution
             will be approximately  10 mg/mL.

       7.2.3.2 SURROGATE PRIMARY DILUTION STANDARD (20 |ig/mL) -
             Prepare a primary dilution standard (PDS) at a concentration of 20 |ig/mL
             (or other suitable concentration) by adding 50 jiL (or suitable volume) of
             the SUR stock standard to a  25-mL volumetric flask containing MTBE.
             Dilute the flask to volume, and mix thoroughly.  Transfer the SUR PDS to
             an amber glass vial with a PTFE-lined screw cap and store at < 0 °C.
             Addition of 20 jiL of the primary dilution standard to the 40-mL aqueous
             sample results in a surrogate concentration of  10 ng/mL.

7.2.4   ANALYTE STANDARD SOLUTIONS - Obtain the analytes listed in the table
       in Section 1.1 as neat or solid  free acid standards or as commercially prepared
       ampulized solutions from a reputable standard manufacturer.  The use of
       premethylated standards is not allowed for the preparation of analyte standards.
       Prepare the Analyte Stock and Primary Dilution Standards as described below.

       7.2.4.1 ANALYTE STOCK SOLUTION - Prepare separate stock standard
             solutions for each analyte of interest at a concentration of 1-5 mg/mL in
             MTBE. Method analytes are available as neat  materials or ampulized
             solutions (> 99% purity) from a number of commercial suppliers.

             7.2.4.1.1  For analytes which are solids in their pure form, prepare stock
                       standard solutions by accurately weighing approximately 0.01
                       to 0.05 grams of  pure material in a 10-mL volumetric flask
                       using the technique described in  Section 7.2.1.1.
                             552.3-14

-------
             7.2.4.1.2  For analytes which are liquid in their pure form at room
                       temperature, add 10.0 jiL of standard material to the flask using
                       the technique described in Section 7.2.1.2.

       7.2.4.2 ANALYTE PRIMARY DILUTION STANDARD (PDS)- Prepare the
             Analyte PDS solution by combining and diluting the Analyte Stock
             Solutions prepared above (Sect. 7.2.4.1) with MTBE.  This solution will
             be used to prepare the Analyte Secondary Dilution Standard. As a
             guideline to the analyst, the concentration used in the primary dilution
             standard solution during method development was 100 |ig/mL for all ten
             analytes.

       7.2.4.3 ANALYTE SECONDARY DILUTION STANDARD (SDS)- The Analyte
             SDS is used to fortify reagent water for calibration standards. Prepare at
             least one Analyte SDS by diluting the Analyte PDS with methanol. Two
             Analyte SDSs were used during method development.  The first was
             prepared at a concentration of 5.00 |ig/mL (1/20 dilution of the Analyte
             PDS) and the second at a concentration of 16.7 |ig/L (1/6 dilution of the
             Analyte PDS).  The SDS should be made daily.  It should not be stored,
             since methanol  will derivatize the acids that it contains. The lowest
             calibration standard concentration must be at or below the MRL of each
             analyte. The concentrations of the other standards should  span the range of
             concentration expected in the Field Samples or the working range of the
             instrument.

7.2.5   CALIBRATION STANDARDS - At least 5 calibration standards are required to
       prepare the initial calibration curve (Sect. 10.2). Fortify an appropriate number of
       reagent water solutions with varying amounts of the Analyte PDS over the
       concentration range of interest. During method development, the Calibration
       Standards typically ranged from 1.0 |ig/L to 20 |ig/L. The lowest standard must
       be at or below the MRL, which will depend upon instrument sensitivity. Because
       this method employs a  procedural calibration technique , these standards must be
       treated like samples and require the addition of all preservation and other
       reagents. They are extracted by the procedure described in Section 11.
                             552.3-15

-------
8.      SAMPLE COLLECTION. PRESERVATION. AND STORAGE

       8.1    SAMPLE BOTTLE PREPARATION

             8.1.1   Grab samples must be collected in accordance with conventional sampling
                    practices(7) using amber glass containers with PTFE-lined screw caps and
                    capacities of at least 50 mL.

             8.1.2   Prior to shipment to the field,  add crystalline or granular ammonium chloride
                    (NH4C1) to the  sample containers to produce a concentration of 100 mg/L in the
                    Field Sample. For a typical 50-mL sample, this requires 5 mg of ammonium
                    chloride.

                    Note: Enough ammonium chloride must be added to the sample to convert the
                    free chlorine residual in the sample matrix to combined chlorine. Chloramines,
                    formed by the reaction of hypochlorite with the ammonium ion, do not react
                    further to produce additional haloacetic acids at significant concentrations and
                    protect against  microbiological degradation.(8)  This  concentration of ammonium
                    chloride was determined to convert 8 mg/L of free chlorine residual to combined
                    chlorine.

       8.2.   SAMPLE COLLECTION

              8.2.1  Fill sample bottles but take care not to flush out the ammonium chloride. Because
                    the target analytes of this method are not volatile, it is not necessary to ensure that
                    the sample bottles are completely headspace free.

              8.2.2  When sampling from a water tap, remove the aerator. Open the tap and allow the
                    system to flush until the water temperature has stabilized (usually about 3-5
                    minutes). Collect samples from the flowing system.

             8.2.3   When sampling from an open body of water, fill a 1-quart  wide-mouth bottle or
                    1 liter beaker with sample from a representative area, and carefully fill  sample
                    vials from the container.

             8.2.4   After collecting the sample, seal the bottle and agitate by hand for 15 seconds.
                                          552.3-16

-------
       8.3    SAMPLE SHIPMENT AND STORAGE - All samples must be chilled during shipment
             and must not exceed 10 °C during the first 48 hours after collection. Samples must be
             confirmed to be at or below 10 °C when they are received at the laboratory.  Samples
             stored in the lab must be held at or below 6 °C and protected from light until extraction.
             Samples should not be frozen.

       8.4    SAMPLE AND EXTRACT HOLDING TIMES - Chlorinated Field Samples that are
             preserved according to the method should not exhibit biological degradation of analytes
             during the allotted  14-day storage time. Samples must be extracted within 14 days.
             Unchlorinated Field Samples should be extracted as soon as possible to prevent
             biological degradation of analytes. Extracts must be stored at < -10 °C or less and
             protected from light in glass vials with PTFE-lined caps.  MTBE extracts must be
             analyzed within 21 days of extraction. TAME extracts must be analyzed within 28 days
             of extraction.

             8.4.1   The brominated trihaloacetic acids  can exhibit degradation during storage.
                    Tribromoacetic acid (TBAA), the least stable HAA ester, is degraded to
                    bromoform. This is thought to occur as a result of peroxide contamination in the
                    solvent.   Low concentrations of peroxides that were barely detectable with the
                    commercially available peroxide test strips were found to generate significant
                    levels of bromoform in standard solutions.  Bromoform can be chromatographed
                    under method conditions (see Tables 1-4), and can be monitored as an indication
                    of high peroxide levels in the solvent. If large bromoform peaks are observed in
                    HAA-fortified  reagent water samples (or calibration standards), the ether solvent
                    should be purified or replaced. Storing these solvents under nitrogen minimizes
                    peroxide formation.

             8.4.2   Extracts that were prepared using the TAME procedure showed much lower
                    tendency to form bromoform during the method development studies. This is
                    thought to be associated with lower levels of peroxides.  This may be due to
                    TAME being stored over molecular sieves (Sect. 7.1.3).
9.      QUALITY CONTROL

       9.1    Each laboratory using this method is required to operate a formal Quality Control (QC)
             program. The requirements of this program consist of an Initial Demonstration of
             Capability (IDC), and subsequent analysis in each analysis batch of a Laboratory

                                           552.3-17

-------
      Reagent Blank (LRB), Continuing Calibration Check Standards (CCCs), a Laboratory
      Fortified Sample Matrix (LFSM), and either a Laboratory Fortified Sample Matrix
      Duplicate (LFSMD) or a Field Duplicate Sample.  This section details the specific
      requirements for each QC parameter.  The QC criteria discussed in the following sections
      are summarized in Section 17, Tables 15 and 16.  These criteria are considered the
      minimum acceptable QC criteria, and laboratories are encouraged to institute additional
      QC practices to meet their specific needs.

      9.1.1  Process all quality control samples through all steps of Section 11, including
             methylation.  Sample preservatives as described in Section 8.1 must be added
             prior to extracting and analyzing the quality control samples.

9.2   INITIAL DEMONSTRATION OF CAPABILITY (IDC) - Requirements for the Initial
      Demonstration of Capability are described in the following sections and summarized in
      Section 17, Table 15.

      9.2.1  INITIAL DEMONSTRATION OF LOW SYSTEM BACKGROUND - Before
             any samples are analyzed, it must be demonstrated that a laboratory reagent blank
             (LRB) is reasonably free of contamination.

      9.2.2  INITIAL DEMONSTRATION OF ACCURACY - Prior to the analysis of the
             IDC samples, verify calibration accuracy with the preparation and analysis of a
             mid-level  QCS as defined in Section 9.10.  If the analyte recovery is not + 30% of
             the expected value, the accuracy of the method is unacceptable. The source of the
             problem must be identified and corrected.  After the accuracy of the calibration
             has been verified, prepare and analyze 4-7 replicate LFBs (or CCCs in this
             method) fortified at 10 |ig/L, or near the mid-range of the initial calibration curve,
             according to the procedure described in Section 11.  Sample preservatives as
             described  in Section 8.1.2 must also be added to these samples. The average
             recovery of the replicate values must be within ± 20% of the expected value.

      9.2.3   INITIAL DEMONSTRATION OF PRECISION - Using the same set of replicate
             data generated for Section 9.2.2, calculate the standard deviation and relative
             standard deviation of the replicate recoveries. The relative standard deviation
             (RSD) of the results of the replicate analyses must be less than 20%.

      9.2.4  DETECTION LIMIT (DL) DETERMINATION - Prepare, extract and analyze at
             least seven replicate LFBs at a concentration estimated to be near the Detection

                                    552.3-18

-------
       Limit over a period of at least three days (both extraction and analysis should be
       conducted over at least three days) using the procedure described in Section 11.
       The fortification level may be estimated by selecting a concentration with a signal
       of 2 to 5 times the noise level. The appropriate concentration will be dependent
       upon the sensitivity of the GC/ECD system being used. Sample preservatives as
       described in Section 8.1.2 must be added to these samples. Calculate the
       Detection Limit using the equation

                         DL — St(n _ i? i _a = 0.99)

       where
             V -1, i - a = o 99) = Students t value for the 99% confidence level with n-1
                  degrees of freedom ( 3.143 for 7 replicates)
             n = number of replicates, and
             S = standard deviation of the replicate analyses.

       NOTE: Do not subtract blank values when performing DL calculations.  The DL
       is a statistical determination of precision only.(1)  If the DL replicates are fortified
       at a low enough concentration, it is likely that they will not meet precision and
       accuracy criteria, and may result in a calculated DL that is higher than the fortified
       concentration. Therefore no precision and accuracy criteria are specified for the
       DL.

9.2.5   Minimum Reporting Level (MRL) - The MRL is the threshold concentration of
       an analyte that a laboratory can expect to accurately quantitate in an unknown
       sample.  The MRL should be established at an analyte concentration that is at
       least 3 times the DL or at a concentration which would yield a signal-to-noise
       (S/N) ratio of greater than or equal to five.  Depending upon the study's data
       quality objectives it may be set at a higher concentration.  The concentration of
       the lowest calibration standard must be at or below the MRL.

9.2.6   METHOD MODIFICATIONS  - The analyst is permitted to modify GC columns,
       GC conditions (see Tables 1-4), and internal standards (see Sect. 7.2.2) or
       surrogate standards (see  Sect. 7.2.3). Other detectors (see Sect. 6.14) may be used
       if they have equivalent or better selectivity and have sufficient sensitivity. Each
       time such method modifications are made, the analyst must repeat the procedures
       of the IDC (Sect. 9.2).
                              552.3-19

-------
9.3    LABORATORY REAGENT BLANK (LRB) - An LRB is required with each extraction
       batch (Sect. 3.1) of samples to determine the background system contamination. If the
       LRB produces a peak within the retention time window of any analyte that would prevent
       the determination of that analyte, determine the source of contamination and eliminate the
       interference before processing samples. Background contamination must be reduced to
       an acceptable level before proceeding. Background from method analytes or
       contaminants that interfere with the measurement of method analytes must be below 1/3
       of the MRL.  If the target analytes are detected in the LRB at concentrations equal to or
       greater than this level, then all data for the problem analyte(s) must be considered invalid
       for all samples in the extraction batch.

9.4    CONTINUING CALIBRATION CHECK (CCC) - CCC Standards are prepared in the
       same extraction batch as the samples of interest. They must contain all compounds of
       interest,  and they are extracted in the same manner as the Field Samples and calibration
       solutions used to prepare the initial calibration curve.  Calibration checks, prepared with
       the samples being analyzed, are required at the beginning of each analysis batch, after
       every ten samples, and at the  end of the analysis batch.  See Section 10.3 for
       concentration requirements, frequency requirements, and acceptance criteria.

9.5    LABORATORY FORTIFIED BLANK (LFB) - Since this method utilizes procedural
       calibration standards, which are fortified reagent waters, there is no difference between
       the LFB and the continuing calibration check standard.  Consequently, the analysis of an
       LFB is not required; however, the acronym LFB is used for clarity in the IDC.

9.6    INTERNAL STANDARDS (IS) - The analyst must monitor the peak area of each
       internal  standard in all injections during each analysis day.  The IS response (as indicated
       by peak  area) for any chromatographic run must not deviate by more than ± 50% from the
       average  area measured during the initial calibration for that IS.  A poor injection could
       cause the IS area to exceed these criteria.  Inject a second aliquot of the suspect extract to
       determine whether the failure is due to poor injection or instrument response drift.

       9.6.1  If the reinjected aliquot produces an acceptable internal standard response, report
             results for that aliquot.

       9.6.2  If the internal standard area for the reinjected extract deviates greater than 50%
             from  the initial calibration average, the analyst should check the continuing
             calibration check standards that ran before and after the sample. If the continuing
             calibration check fails the criteria of Section 10.3, recalibration is in order per

                                    552.3-20

-------
              Section 10. If the calibration standard is acceptable, extraction of the sample
              should be repeated provided the sample is still within holding time. Otherwise,
              report results obtained from the reinjected extract, but annotate as suspect.

9.7    SURROGATE RECOVERY - The surrogate standard is fortified into the aqueous
       portion of all samples, LRBs,  CCCs, LFSMs, and LFSMDs prior to extraction. It is also
       added to the calibration standards. The surrogate is a means of assessing method
       performance from extraction to final chromatographic measurement. Calculate the
       recovery (R) for the surrogate using the equation


                                     Ml
                                 R=  —  xlQO%

       where
              A = calculated surrogate concentration for the QC or Field Sample, and
              B = fortified concentration of the surrogate.

       9.7.1   When surrogate recovery from a sample, blank,  or CCC is less than 70% or
              greater than 130%, check (1) calculations to locate possible errors, (2) standard
              solutions for degradation, (3) contamination, and (4) instrument performance.
              Correct the problem and reanalyze the extract.

       9.7.2   If the extract reanalysis meets the surrogate recovery criterion, report only data for
              the reanalyzed extract.

       9.7.3   If the extract reanalysis fails the 70-130% recovery criterion, the analyst should
              check the calibration by injecting the last calibration standard that passed. If the
              calibration standard fails the criteria of Section 9.7.1, recalibration is in order per
              Section 10.2.  If the calibration standard is acceptable, extraction  of the sample
              should be repeated provided the sample is still within the holding time.  If the re-
              extracted sample also fails the recovery criterion, report all data for that sample as
              suspect/surrogate recovery to inform the data user that the results are suspect due
              to surrogate recovery.

9.8    LABORATORY FORTIFIED SAMPLE MATRIX (LFSM) - Analysis of an LFSM
       (Sect. 3.7) is required in each  extraction batch and is used to determine that the sample
       matrix does not adversely affect method accuracy.  If the occurrence of target analytes in
       the samples is infrequent, or if historical trends are unavailable, a second LFSM or

                                     552.3-21

-------
LFSMD must be prepared, extracted, and analyzed from a duplicate field sample used to
prepare the LFSM to assess method precision.  Extraction batches that contain LFSMDs
do not require the analysis of a Field Duplicate (Sect. 9.9).  If a variety of different
sample matrices are analyzed regularly, for example, drinking water from groundwater
and surface water sources, method performance should be established for each.  Over
time, LFSM data should be documented for all routine sample sources for the laboratory.

9.8.1   Within each extraction batch, a minimum of one Field Sample is fortified as an
       LFSM for every 20 samples extracted.  The LFSM is prepared by spiking a
       sample with an appropriate amount of Analyte PDS (Sect. 7.2.4.2). Select a
       spiking concentration that is greater than or equal to the matrix background
       concentration, if known.  Use historical data  and rotate through the designated
       concentrations when selecting a fortifying concentration.

9.8.2   Calculate the recovery (R) for each analyte using the equation

                              (A-B)
                         R=	-x  100%
                                U-
       where
              A = measured concentration in the fortified sample,
              B = measured concentration in the unfortified sample, and
              C = fortification concentration.

9.8.3   Analyte recoveries may exhibit matrix bias.  For samples fortified at or above
       their native concentration, recoveries should  range between 70 and 130%, except
       for low-level fortification near or at the MRL (within a factor of 2-times  the MRL
       concentration) where 50  to 150% recoveries  are acceptable.  If the accuracy of any
       analyte falls outside the designated range, and the laboratory performance for that
       analyte is shown to be in control in the CCCs, the recovery is judged to be matrix
       biased. The result for that analyte  in the unfortified sample is labeled
       suspect/matrix to inform the data user that the results are  suspect due to matrix
       effects.

       9.8.3.1 Because HAAs are disinfection by-products, many Field Samples will
              contain a number of HAAs at varying concentrations. Field Samples that
              have native HAA concentrations above the DL but below the  MRL and
              are fortified at concentrations at or near the MRL  should be corrected for
                              552.3-22

-------
                    the native levels in order to obtain meaningful R values. This is the only
                    permitted use of analyte results below the MRL.

9.9    FIELD DUPLICATE OR LABORATORY FORTIFIED SAMPLE MATRIX
       DUPLICATE (FD or LFSMD) - Within each extraction batch, a minimum of one Field
       Duplicate (FD) or Laboratory Fortified Sample Matrix Duplicate (LFSMD) must be
       analyzed. Duplicates check the precision associated with sample collection, preservation,
       storage, and laboratory procedures.  HAAs are typically found in waters disinfected with
       chlorine; however, if target analytes are not routinely observed in Field Samples, an
       LFSMD should be periodically analyzed rather than an FD.

       9.9.1  Calculate the relative percent difference (RPD) for duplicate measurements (FD1
             and FD2) using the equation

                                  \FD\ - FD2\
                          RPD = J	 I  xiQO%
                                (FDl + FD2)/2

             9.9.1.1 RPDs for Field Duplicates should fall in the range of ± 30%.  Greater
                    variability may be observed when Field Duplicates have analyte
                    concentrations that are within a factor of 2 of the MRL.  At these
                    concentrations Field Duplicates should have RPDs that fall in the range of
                    ± 50%.  If the RPD of any analyte falls outside the designated range, and
                    the laboratory performance for that analyte is shown to be in control in the
                    CCC, the recovery is judged to be matrix biased.  The result for that
                    analyte in the unfortified sample is labeled suspect/matrix to inform the
                    data user that the results are suspect due to matrix effects.

       9.9.2  If an LFSMD is analyzed instead of a Field Duplicate, calculate the relative
             percent difference (RPD) for duplicate LFSMs  (LFSM and LFSMD) using the
              equation

                                  LLF.SM - LFSMD]
                           RPD = -J:	i  xioo%
                                 (LFSM + LFSMD)? 2


             9.9.2.1 RPDs for duplicate LFSMs should fall in the range of ± 30% for samples
                    fortified at or above their native concentration. Greater variability may be
                    observed when LFSMs are fortified at analyte concentrations that are
                    within a factor of 2 of the MRL.  LFSMs fortified at these concentrations

                                    552.3-23

-------
                           should have  RPDs that fall in the range of ± 50% for samples fortified at
                           or above their native concentration.  If the RPD of any analyte falls outside
                           the designated range, and the laboratory performance for that analyte is
                           shown to be in control in the CCC, the recovery is judged to be matrix
                           biased.  The result for that analyte in the unfortified sample is labeled
                           suspect/matrix to inform the data user that the results are suspect due to
                           matrix effects.

       9.10   QUALITY CONTROL SAMPLES (QCS) - As part of the IDC (Sect.  9.2), each time a
             new Analyte PDS (Sect. 7.2.4.2) is prepared, and at least quarterly, analyze a QCS
             sample from a source different from the source of the calibration standards. If a second
             vendor is not available  then a different lot of the  standard should be used. The QCS
             should be prepared and analyzed just like a CCC. Acceptance criteria for the QCS is
             identical to the CCCs; the calculated amount for each analyte must be + 30% of the
             expected value.  If measured analyte concentrations are not of acceptable accuracy, check
             the entire analytical procedure to locate and correct the problem.
10.     CALIBRATION AND STANDARDIZATION

       10.1   An acceptable initial calibration must be established during the IDC and prior to
             analyzing Field or QC Samples. After initial calibration is successful, a Continuing
             Calibration Check (CCC) is required at the beginning and the end of each analysis batch,
             and after every tenth sample (Sect. 10.3). Because this is a procedural standard method,
             the analyst will need to make a decision to include either an appropriate number of CCCs
             or an entire initial calibration curve with each extraction batch.  Initial calibration must be
             repeated each time a major instrument modification is made or maintenance is performed.
             Failure to meet CCC criteria may also require recalibration.

       10.2   INITIAL CALIBRATION - This method uses the procedural calibration technique to
             compensate for incomplete methylation of some of the target compounds.  This is most
             pronounced for the brominated trihaloacetic acids. Many of the QC criteria throughout
             this method are expressed in terms of percent recovery.  It should be noted that these
             recoveries are relative to the initial procedural curve rather than absolute recoveries.

             10.2.1  Establish GC operating parameters equivalent to the suggested specifications in
                    Section 17, Table 1 (or Table 2). The GC system must be calibrated using the
                                           552.3-24

-------
       internal standard (IS) technique. Other columns or conditions may be used if
       equivalent or better performance can be demonstrated.

10.2.2  Prepare a set of at least 5 calibration standards as described in Section 7.2.5.  The
       lowest concentration calibration standard must be at or below the MRL, which
       will depend on system sensitivity.  It is recommended that at least four of the
       calibration standards be at concentrations greater than the MRL.

10.2.3  CALIBRATION - Use the GC data system software to generate  a linear
       regression or quadratic calibration curve using the internal standard. The analyst
       may choose whether or not to force zero to obtain a curve that best fits the  data.
       Examples of common GC system calibration curve options are:  1) Ax /A1S  versus
       Qx /Qis; and 2) RRF  versus Ax /A1S.  Establish a relative response  factor by using
       the equation
       where
              Ax = integrated peak area of the analyte,
              A1S = integrated peak area of the internal standard,
              Qx = quantity of analyte injected in ng or concentration units,
              Q1S = quantity of internal standard injected in ng or concentration
                 units, and
              RRF = relative response factor.

10.2.4  Acceptance criteria for the calibration of each analyte is determined by calculating
       the concentration of each analyte and surrogate in each of the analyses used to
       generate the calibration curve.  Each calibration point, except the lowest point, for
       each analyte should calculate to be 70-130% of its expected value. The lowest
       point should calculate to be 50-150% of its expected value. Laboratories that
       have difficulty achieving these criteria will have trouble meeting the QC
       requirements  summarized in Section 10.3.2.  These laboratories should reanalyze
       the calibration standards, restrict the range of calibration, or select an alternate
       method of calibration.
                              552.3-25

-------
       10.3   CONTINUING CALIBRATION CHECK (CCC) - An appropriate number of CCCs
             must be prepared with each extraction batch.  The CCC verifies the initial calibration at
             the beginning and the end of each analysis batch, and after every 10th sample during
             analyses. In this context, a "sample" is considered to be a Field Sample. LFSMs,
             LFSMDs, FSDs, and CCCs are not counted as samples. The beginning CCC for each
             analysis batch must be at or below the MRL in order to verify instrument sensitivity prior
             to any analyses. If standards have been prepared such that all analytes are not in the same
             Calibration Standard,  it may be necessary to analyze two Calibration Standard solutions
             to meet this requirement. Subsequent CCCs should alternate between a medium and a
             high concentration standard.

             10.3.1  Inject  an aliquot of the appropriate concentration calibration check standard
                    solution prepared with the extraction batch and analyze using the same conditions
                    used during the initial calibration.

             10.3.2  Calculate the concentration of each analyte and surrogate in the check standard.
                    The calculated amount for each analyte for medium and high level CCCs must be
                    ± 30% of the expected value.  The calculated amount for the lowest calibration
                    level for each  analyte must be within ± 50% of the expected value.  If these
                    conditions do not exist, then all data for the problem analyte must be considered
                    invalid, and remedial action should be taken which may require recalibration.
                    Any Field Sample extracts that have been analyzed since the last acceptable
                    calibration verification  should be reanalyzed after adequate calibration has been
                    restored. For Extraction Batches with more than 10 Field Samples the analyst
                    may either extract and analyze a third  CCC or reanalyze the mid-level or high-
                    level CCC as the final CCC.
11.    PROCEDURE

      11.1   SAMPLE EXTRACTION

             11.1.1  Remove the samples from storage (Sect. 8.3) and allow them to equilibrate to
                    room temperature.

             11.1.2  Place 40 mL of the water sample into a precleaned 60-mL glass vial with a PTFE-
                    lined screw cap using a clean, graduated cylinder for each sample.
                                          552.3-26

-------
       11.1.3 Add 20 jiL of surrogate standard (20 |ig/mL of 2-bromobutanoic acid in MTBE
             per Section 7.2.3.2) to the aqueous sample.

             NOTE: When fortifying an aqueous sample with components that are contained in
             MTBE solution, be sure that the needle of the syringe is well below the surface of
             the water. After injection, cap the sample and invert once.  This will insure that
             the standard solution is mixed well.

       11.1.4 Adjust the pH to less than or equal to 0.5 by adding up to 2 mL of concentrated
             sulfuric acid. Cap, mix and then check the pH with narrow range pH paper (Sect.
             611) Substitution of other acids is not allowed.

       11.1.5 Add approximately 18 g of muffled sodium sulfate (Sect. 7.1.4) and immediately
             shake until almost all is dissolved. Substitution of other salts for sodium
             sulfate is not allowed. Sodium sulfate is added to increase the ionic strength of
             the aqueous phase and thus further drive the haloacetic acids into the organic
             phase. The addition of salt also decreases the solubility of MTBE or TAME in
             the aqueous phase and allows greater volumetric recovery of the extraction
             solvent.  The addition of this salt should be done immediately after the addition of
             the sulfuric acid so that the heat generated from the addition of the acid (Sect.
             11.1.4) will help dissolve the salt.

       11.1.6 Add exactly 4.0 mL MTBE with IS or TAME with IS (Sect. 7.2.2.3) and shake
             vigorously for three minutes by hand. This may be accomplished for a entire
             Extraction Batch using a test tube rack.

       11.1.7 Allow the phases to separate for approximately 5 minutes.

11.2   SAMPLE METHYLATION WITH ACIDIC METHANOL

       11.2.1 Using a pasteur pipet, transfer 3 mL of the upper MTBE or TAME layer to a
             15-mL graduated conical centrifuge tube.

       11.2.2 Add 3 mL of 10% sulfuric acid in methanol (Sect. 7.1.12) to each centrifuge tube.
             Cap the tube.

       11.2.3 Methylation of the method analytes is accomplished during this step. Careful
             control of both reaction time and reaction temperature are critical to method

                                    552.3-27

-------
      precision and accuracy.(9) Place the tubes in a heating block (or sand bath) at
      50 ± 2 °C (for MTBE) or 60 ± 2 °C (for TAME) and heat for 2 hours (± 10 min).
      The tubes should fit snugly into the heating block to ensure proper heat transfer.
      Verify the reaction temperature by placing a thermometer into a tube containing
      water rather than inserting it into the block to ensure an accurate reading. Placing
      the thermometer directly in the heating block well or sand bath will give a higher
      temperature reading than the actual sample temperature. The MTBE reaction
      temperature is set at its highest practical limit (50 °C), since MTBE boils at 55 °C.
      Similarly, the TAME reaction temperature has been set at 60 °C, since methanol
      boils at 65 °C.  This helps prevent solvent loss from the reaction tubes.

      11.2.3.1 Even at 2 hours, methylation for some of the more sterically hindered
               compounds like TBAA, CDBAA and BDCAA is not complete.
               Shortening reaction time will decrease methylation efficiencies for
               compounds and result in lower precision.

      11.2.3.2 Methylation efficiencies are increased with increasing temperature (and
               time). The TAME procedure has higher methylation efficiencies than
               MTBE procedure for the sterically hindered HAAs like the brominated
               trihaloacetic acids (Sect. 11.2.3.1) and is more precise for these targets.
               Methylation temperature, however, should not be increased above
               the recommended reaction temperatures (Sect. 11.2.3).

      11.2.3.3 Lower reaction temperatures or times are not allowed. Care should
               be taken to ensure that the calibration standards are heated identically to
               Field Samples in the extraction batch.

      11.2.3.4 Methylation may be accomplished by heating the reaction tubes with a
               water bath, provided the water bath is not covered in a way as to cause
               the entire tube to be heated. If tube walls are heated, the tubes can lose
               some of their contents leading to higher variability in analytical results.
               A water bath that is covered with a layer of small, floating plastic
               spheres may be used (Sect. 6.10).

11.2.4 Remove the centrifuge tubes from the heating source and allow them to cool
      before removing their caps.
                             552.3-28

-------
       11.2.5 Add 7 mL of a 150 g/L sodium sulfate solution (Sect. 7.1.10) to each centrifuge
             tube. Vortex each tube to ensure full equilibration between the phases. Allow the
             two phases to settle fully, but do not allow the tubes to sit more than a few
             minutes. Adding the sodium sulfate solution may cause some loss of the formed
             HAA-esters through acid-catalyzed hydrolysis over prolonged periods.

       11.2.6 Remove and discard the lower (acidic aqueous methanol) phase from each tube
             with a long pasteur pipet. Leave no more than 0.3 mL of aqueous phase to ensure
             complete neutralization in the following step.

       11.2.7 Add 1 mL of saturated sodium bicarbonate solution (Sect. 7.1.11). Vortex each
             centrifuge tube for several seconds at least four times to complete the
             neutralization reaction. Loosen the tube caps after the first agitation to release the
             evolved CO2.

       11.2.8 Transfer 1 mL of the upper ether layer to an auto-sampler vial. A duplicate vial
             should be filled using the excess extract.

       11.2.9 Analyze the samples as soon as  possible.  Store the extracts at <-10 °C (Sect. 8.4).

11.3    GAS CHROMATOGRAPHY

       11.3.1 The instrument used in the development of this method was equipped with a low
             volume (150 |jL) micro electron capture detector (BCD). Other configurations are
             allowed as described in Section 6.14.

       11.3.2 COLUMN SELECTION AND INSTALLATION - Strict attention should be paid
             to established column installation guidelines with regard to the proper cutting and
             placement of the capillary columns within the instrument. The brominated
             trihaloacetic acids are particularly  sensitive to the condition  of the injection port
             and GC column. If the response for these or other method analytes diminish,
             trimming approximately 0.5 m  from the head of the column and replacing the GC
             inlet liner often restores the response for these analytes. Quartz inlet liners and
             inlet liners with standard deactivation were found to require less frequent
             maintenance than those with Siltek™ deactivation for this method. If conditions
             in the laboratory necessitate frequent column trimming, a guard column is
             recommended.
                                    552.3-29

-------
              11.3.3 Method development was conducted with the GC set in constant pressure mode.
                    If constant velocity mode is used by the analyst, linear velocities should be
                    reduced from those listed in Tables 1-4 to obtain similar resolution.

       11.4    ANALYSIS OF EXTRACTS

              11.4.1 Establish operating conditions as described in Section 17, Table 1  or Table 2
                    (Table 3 or 4 if performing confirmation). Confirm that retention times,
                    compound separation, and resolution are similar to those summarized in Tables 1-
                    4 and Figures 1-4.

              11.4.2 Establish an appropriate relative retention time window for each target and
                    surrogate to identify them in the QC and Field Samples. This should be based on
                    measurements of actual relative retention time variation for each compound in
                    standard solutions analyzed on the GC over the course of time. Plus or minus
                    three times the standard deviation of the relative retention time observed for each
                    compound while establishing the initial calibration can be used to calculate a
                    suggested window size; however, the experience of the analyst should weigh
                    heavily on the determination of the appropriate retention window.

              11.4.3 Calibrate the instrument as described in Section 10.2 or confirm the calibration is
                    still valid by analyzing CCCs as described in Section 10.3. Begin analyzing Field
                    and QC Samples at their appropriate frequency by injecting aliquots under the
                    same  conditions used to establish the initial calibration.

              11.4.4 The analyst must not extrapolate  beyond the established calibration range.  If an
                    analyte result exceeds the range of the  initial calibration curve, the extract may be
                    diluted with MTBE or TAME containing the internal standard (Sect. 7.2.1.3), and
                    the diluted extract injected. Acceptable surrogate performance (Sect. 9.7) should
                    be determined from the undiluted sample extract.  Incorporate the dilution factor
                    into final concentration calculations. The dilution will also affect analyte MRLs.
12.    DATA ANALYSIS AND CALCULATION

       12.1    Identify the method analytes in the sample chromatogram by comparing the retention
              time of the suspect peak to retention time of an analyte peak in a calibration standard.


                                           552.3-30

-------
             Surrogate retention times should be confirmed to be within acceptance limits (Sect.
             11.4.2) even if no target compounds are detected.

       12.2   Calculate the analyte concentrations using the initial calibration curve generated as
             described in Section 10.2.  Quantitate only those values that fall between the MRL and
             the highest calibration standard. Samples with target analyte responses that exceed the
             highest standard require dilution and reanalysis (Sect. 11.4.4).

       12.3   Analyte identifications should be confirmed using the confirmation column specified in
             Table 3 (or Table 4) or another column that is dissimilar to the primary column. GC/MS
             confirmation is acceptable if the analyte concentrations are sufficient.

       12.4   Adjust the calculated concentrations of the detected analytes to reflect the initial sample
             volume and any dilutions performed.

       12.5   Analyte concentrations are reported in jig/L as the total free acid (usually to 2 significant
             figures); however, calculations should use all available digits of precision.
13.    METHOD PERFORMANCE

       13.1   PRECISION, ACCURACY, AND DETECTION LIMITS - Tables for these data are
             presented in Section 17. Detection Limits are presented in Table 5 and were calculated
             using the formula presented in Section 9.2.4. Single laboratory precision and accuracy
             data are presented in Tables 6-11.

       13.2   SAMPLE STORAGE STABILITY STUDIES - An analyte storage stability study was
             conducted by fortifying the analytes (10 jig/L of each analyte) into a chlorinated surface
             water that was collected, preserved, and stored as described in Section 8. The average of
             triplicate analyses, conducted on days 0, 3, 7, and 14 are presented in Table 12.

       13.3   EXTRACT STORAGE STABILITY STUDIES - Extract storage stability studies were
             conducted on TAME and MTBE extracts obtained from a chlorinated surface water
             fortified at 10 |ig/L. The average of triplicate injections are reported in Tables 13 and 14.
             Extract storage stability can decline in the presence of peroxides. This is discussed in
             Section 8.4.
                                           552.3-31

-------
14.     POLLUTION PREVENTION

       14.1   This method utilizes liquid-liquid microextraction to extract analytes from water.  It
             requires the use of very small volumes of organic solvent and very small quantities of
             pure analytes, thereby minimizing the potential hazards to both the analyst and the
             environment as compared to the use of large volumes of organic solvents in conventional
             liquid-liquid extractions.

       14.2   For information about pollution prevention that may be applicable to laboratory
             operations, consult "Less is Better: Laboratory Chemical Management for Waste
             Reduction" available from the American Chemical Society's Department of Government
             Relations and Science Policy,  1155 16th Street N.W., Washington, D.C., 20036.
15.    WASTE MANAGEMENT

       15.1   The analytical procedures described in this method generate relatively small amounts of
             waste since only small amounts of reagents and solvents are used. The matrices of
             concern are finished drinking water or source water.  However, the Agency requires that
             laboratory waste management practices be conducted consistent with all applicable rules
             and regulations, and that laboratories protect the air, water, and land by minimizing and
             controlling all releases from fume hoods and bench operations. Also, compliance is
             required with any sewage discharge permits and regulations, particularly the hazardous
             waste identification rules and land disposal restrictions. For further information on waste
             management, consult "The Waste Management Manual for Laboratory Personnel" also
             available from the American Chemical Society at the address in Section 14.2.
16.    REFERENCES

       1.     Glaser, J.A., Foerst, D.L., McKee, G.D., Quave, S.A., and Budde, W.L., "Trace Analyses
             for Wastewaters." Environ. Sci. Technol.. 15 (1981) 1426-1435.

       2.     Giam, C.S., Chan, H.S., and Nef, G.S., "Sensitive Method for Determination of Phthalate
             Ester Plasticizers in Open-Ocean Biota Samples," Analytical Chemistry. 47, 2225 (1975).
                                           552.3-32

-------
3.      Giam, C.S., and Chan, H.S., "Control of Blanks in the Analysis of Phthalates in Air and
       Ocean Biota Samples," U.S. National Bureau of Standards, Special Publication 442,
       701-708, (1976).

4.      "OSHA Safety and Health Standards, General Industry," (29CRF1910). Occupational
       Safety and Health Administration, OSHA 2206, (Revised, Jan. 1976).

5.      "Carcinogens-Working with Carcinogens," Publication No. 77-206, Department of
       Health, Education, and Welfare, Public Health Service, Center for Disease Control,
       National Institute of Occupational Safety and Health, Atlanta, Georgia, August 1977.

6.      "Safety In Academic Chemistry Laboratories," 3rd Edition, American Chemical Society
       Publication, Committee on Chemical Safety, Washington, D.C., (1979).

7.      ASTM Annual Book of Standards, Part II, Volume 11.01, D3370-82, "Standard Practice
       for Sampling Water," American Society for Testing and Materials, Philadelphia, PA,
       1986.

8.      Pepich, B.V., Domino, M.M., Munch, D.J. and Fair, P.S.  "Validating Sample
       Preservation Techniques and Holding Times for the Approved Compliance Monitoring
       Methods for Haloacetic Acids Under the U.S. EPA's Stage 1 D/DBP Rule,"  Submitted to
       Wat. Res., Sept. 2002.

9.      Xie, Y., Rashid, I, Zhou, H., and Gammie, L. "Acidic Methanol Methylation for HAA
       Analysis: Limitations and Possible Solutions" JAWWA, 94:11, 115-122 (Nov. 2002).
                                    552.3-33

-------
17. TABLES. DIAGRAMS. FLOWCHARTS. AND VALIDATION DATA

TABLE 1:    CHROMATOGRAPHIC CONDITIONS AND AVERAGE RETENTION TIME
               DATA FOR THE PRIMARY COLUMN (DB-1701) USING MTBE
Compound
Monochloroacetic acid (MCAA)
Monobromoacetic acid (MBAA)
Dalapon
Dichloroacetic acid (DCAA)
Bromoform**
Trichloroacetic acid (TCAA)
1,2,3 Trichloropropane (IS)
Bromochloroacetic acid (BCAA)
2-bromobutanoic acid (SUR)
Bromodichloroacetic acid (BDCAA)
Dibromoacetic acid (DBAA)
Chlorodibromoacetic acid (CDBAA)
Tribromoacetic acid (TBAA)
Average Tr
(min)*
9.29
14.19
14.69
15.07
17.12
18.89
20.69
21.22
21.63
23.60
24.01
25.55
26.88
RSD
(%)
0.03
0.02
0.02
0.02
0.01
0.01
0.01
0.01
0.01
0.01
0.005
0.004
0.003
*The average retention time represents the average of 8 injections of fortified reagent water extracts that had been fortified at
10 |ig/L of each analyte.
"Bromoform is not a target analyte. RT value is provided for information only.

Primary Column: DB-1701, 30 m x 0.25 mm i.d., 0.25 |im film thickness.

Injector:        Injector temperature, 210 °C; 2-mm straight quartz liner; injection volume 1 |iL; splitless injection hold for
               45 sec then purge @ 30 mL/min.

GC Program:    (for MTBE) 40 °C initial held for 10 minutes, program at 2.5 °C/min to 65 °C, then 10 °C/min to 85 °C, then
               20 °C/min to 205 °C, hold for 7 min.

Detector:        Agilent Micro ECD (150 \iL volume); detector temperature, 290 °C; detector make up gas 95% Ar/5% CH4
               at 20 mL/min.

Carrier Gas:     Helium (UHP), set at constant pressure.  Initial carrier gas velocity (at 40 °C) 33 cm/sec.

Data Collection:  Agilent GC Chemstation with a digitization rate of 20 Hz.
                                                552.3-34

-------
TABLE 2:     CHROMATOGRAPHIC CONDITIONS AND AVERAGE RETENTION TIME
                DATA FOR THE PRIMARY COLUMN (DB-1701) USING TAME
Compound
Monochloroacetic acid (MCA A)
Monobromoacetic acid (MBAA)
Dalapon
Dichloroacetic acid (DCAA)
Bromoform**
Trichloroacetic acid (TCAA)
1,2,3 Trichloropropane (IS)
Bromochloroacetic acid (BCAA)
2-bromobutanoic acid (SUR)
Bromodichloroacetic acid (BDCAA)
Dibromoacetic acid (DBAA)
Chlorodibromoacetic acid (CDBAA)
Tribromoacetic acid (TBAA)
Average Tr
(min)*
5.94
8.76
9.17
9.34
11.06
12.28
13.56
13.87
14.20
15.78
16.14
17.58
18.89
RSD
(%)
0.08
0.04
0.04
0.04
0.02
0.02
0.01
0.01
0.01
0.004
0.003
0.002
0.02
*The average retention time represents the average of 8 injections of fortified reagent water extracts that had been fortified at
10 ng/L of each analyte.
"Bromoform is not a target analyte. RT value is provided for information only.

Primary Column: DB-1701, 30 m x 0.25 mm i.d., 0.25 urn film thickness.

Injector:         Injector temperature, 210 °C; 2-mm straight quartz liner; injection volume 1 |iL; splitless injection hold for
                45 sec then purge @ 30 mL/min.

GC Program:    (for TAME) 55 °C initial hold 8 minutes, program at 2.5 °C/min to 65 °C, then 10 °C/min to 85 °C, then 20
                °C/min to 205 °C, hold for 7 min.

Detector:        AgilentMicro ECD (150 |iL volume); detector temperature, 290 °C; detectormake up gas 95% Ar/5% CH4
                at 20 mL/min.

Carrier Gas:      Helium (UHP), set at constant pressure. Initial carrier gas velocity (at 55 °C) 32 cm/sec.

Data Collection:  Agilent GC Chemstation with a digitization rate of 20 Hz.
                                                  552.3-35

-------
TABLE 3:    CHROMATOGRAPHIC CONDITIONS AND AVERAGE RETENTION TIME
               DATA FOR THE CONFIRMATION COLUMN (DB-5.625) USING MTBE
Compound
Monochloroacetic acid (MCAA)
Monobromoacetic acid (MBAA)
Dichloroacetic acid (DCAA)
Dalapon
Bromoform**
Trichloroacetic acid (TCAA)
Bromochloroacetic acid (BCAA)
1,2,3 Trichloropropane (IS)
2-bromobutanoic acid (SUR)
Dibromoacetic acid (DBAA)
Bromodichloroacetic acid (BDCAA)
Chlorodibromoacetic acid (CDBAA)
Tribromoacetic acid (TBAA)
Average Tr
(min)*
5.28
8.31
9.05
10.63
12.74
14.73
15.25
15.65
17.41
21.08
21.39
24.35
25.94
RSD
(%)
0.04
0.01
0.01
0.02
0.02
0.02
0.02
0.01
0.02
0.01
0.01
0.002
0.000
*The average retention time represents the average of 8 injections of fortified reagent water extracts that had been fortified at
10 ng/L of each analyte.
"Bromoform is not a target analyte. RT value is provided for information only.

Confirmation
Column:        DB-5.625, 30 m x 0.25 mm i.d., 0.25 urn film thickness.

Injector:        Injector temperature, 210 °C; 2-mm straight quartz liner; injection volume 1 \iL ; splitless injection hold for
               45 sec then purge @ 30 mL/min.

GC Program:    (for MTBE) 40 °C initial held for 10 minutes, program at 2.5 °C/min to 65 °C, then 10 °C/min to 85 °C, then
               20 °C/min to 205 °C,. Post run 210 °C hold for 7 min.

Detector:        Agilent Micro ECD (150 |iL volume); detector temperature, 290 °C; detector make up gas 95% Ar/5% CH4
               at 20 mL/min.

Carrier Gas:     Helium (UHP), set at constant pressure. Initial carrier gas velocity (at 40 °C) 32 cm/sec.

Data Collection: Agilent GC Chemstation with a digitization rate of 20 Hz.
                                                 552.3-36

-------
TABLE 4:     CHROMATOGRAPHIC CONDITIONS AND AVERAGE RETENTION TIME
                DATA FOR THE CONFIRMATION COLUMN (DB-5.625) USING TAME
Compound
Monochloroacetic acid (MCAA)
Monobromoacetic acid (MBAA)
Dichloroacetic acid (DCAA)
Dalapon
Bromoform**
Trichloroacetic acid (TCAA)
Bromochloroacetic acid (BCAA)
1,2,3 Trichloropropane (IS)
2-bromobutanoic acid (SUR)
Dibromoacetic acid (DBAA)
Bromodichloroacetic acid (BDCAA)
Chlorodibromoacetic acid (CDBAA)
Tribromoacetic acid (TBAA)
Average Tr
(min)*
4.07
5.47
5.81
6.61
7.97
9.11
9.44
9.79
11.05
13.77
14.01
16.45
17.96
RSD
(%)
0.26
0.14
0.12
0.10
0.07
0.06
0.05
0.05
0.03
0.01
0.01
0.008
0.004
*The average retention time represents the average of 8 injections of fortified reagent water extracts that had been fortified at
10 ng/L of each analyte.
"Bromoform is not a target analyte. RT value is provided for information only.

Confirmation
Column:         DB-5.625, 30 m x 0.25 mm i.d., 0.25  |im film thickness.

Injector:         Injector temperature, 210 °C; 2-mm straight quartz liner; injection volume 1 |iL; splitless injection hold for
                45 sec then purge @ 30 mL/min.

GC Program:    (for TAME) 55 °C initial held for 8 minutes, program at 2.5 °C/min to 65 °C, then 10 °C/min to 85 °C, then
                20 °C/min to 205 °C,. Post run 210 °C hold for 7 min.

Detector:        AgilentMicro ECD (150 |iL volume); detector temperature, 290 °C; detectormake up gas 95% Ar/5% CH4
                at 20 mL/min.

Carrier Gas:      Helium (UHP), set at constant pressure. Initial carrier gas velocity (at 55 °C) 32 cm/sec.

Data Collection:  Agilent GC Chemstation with a digitization rate of 20 Hz.
                                                  552.3-37

-------
TABLE 5:    DETECTION LIMITS IN REAGENT WATER USING THE MTBE AND TAME
             PROCEDURES
Compound
Monochloroacetic acid (MCAA)
Monobromoacetic acid (MBAA)
Dalapon
Dichloroacetic acid (DCAA)
Trichloroacetic acid (TCAA)
Bromochloroacetic acid (BCAA)
Bromodichloroacetic acid (BDCAA)
Dibromoacetic acid (DBAA)
Chlorodibromoacetic acid (CDBAA)
Tribromoacetic acid (TBAA)
MTBE
Fortification
Level
(ug/L)
1.00
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
Detection
Limit3
(Hg/L)
0.17
0.027
0.024
0.020
0.019
0.016
0.034
0.012
0.054
0.11
TAME
Fortification
Level
(Hg/L)
1.00
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
Detection
Limit8
(Hg/L)
0.20
0.13
0.14
0.084
0.024
0.029
0.031
0.021
0.035
0.097
Tortified reagent waters were extracted and analyzed over 3 days for 7-9 replicates following the
procedure outlined in Section 9.  MTBE detection limits were determined with 9 replicates. TAME
detection limits were determined with 7 replicates.
                                          552.3-38

-------
TABLE 6:    PRECISION AND ACCURACY IN REAGENT WATER FORTIFIED AT 1.0 ug/L
Compound
Monochloroacetic acid (MCA A)
Monobromoacetic acid (MBAA)
Dalapon
Dichloroacetic acid (DCAA)
Trichloroacetic acid (TCAA)
Bromochloroacetic acid (BCAA)
Bromodichloroacetic acid (BDCAA)
Dibromoacetic acid (DBAA)
Chlorodibromoacetic Acid (CDBAA)
Tribromoacetic acid (TBAA)
MTBE
Mean
Recovery (%)
95.8
92.2
97.5
93.8
105
102
117
105
125
128
RSD
(%)
(n=8)
4.0
1.8
1.6
1.6
0.52
0.36
1.2
0.63
2.3
4.1
S/N
Ratio8
12
81
110
190
580
580
1400
1300
650
970
TAME
Mean
Recovery
(%)
81.4
90.7
92.8
97.8
107
103
113
105
112
109
RSD
(%)
(n=8)
5.1
3.7
2.1
2.2
0.90
0.94
1.1
0.86
1.5
1.8
S/N
Ratio3
17
92
150
190
520
740
1800
1800
1600
1100
aSignal-to-noise ratios were calculated for each target compound peak by dividing the peak height for
each compound by the peak-to-peak noise, which was determined for each component from the method
blank over a period of time equal to the full peak width in the target analyte's retention time window.
                                         552.3-39

-------
TABLE 7:   PRECISION AND ACCURACY IN REAGENT WATER FORTIFIED AT 10 ug/L
Compound
Monochloroacetic acid (MCAA)
Monobromoacetic acid (MBAA)
Dalapon
Dichloroacetic acid (DCAA)
Trichloroacetic acid (TCAA)
Bromochloroacetic acid (BCAA)
Bromodichloroacetic acid (BDCAA)
Dibromoacetic acid (DBAA)
Chlorodibromoacetic acid (CDBAA)
Tribromoacetic acid (TBAA)
MTBE
Mean
Recovery"
(%)
101
101
99.8
98.2
102
101
107
102
111
113
RSD
(%)
(n=8)
3.5
2.2
0.88
0.73
1.2
0.72
2.4
0.52
3.8
4.7
TAME
Mean
Recovery"
(%)
102
101
99.9
100
101
101
103
101
104
104
RSD
(%)
(n=8)
2.5
1.6
0.50
0.33
1.1
0.65
1.2
0.59
1.6
2.0
                                   552.3-40

-------
TABLE 8:   PRECISION AND ACCURACY IN CHLORINATED SURFACE WATER
             FORTIFIED AT 1.0 ug/L
Compound
Monochloroacetic acid (MCA A)
Monobromoacetic acid (MBAA)
Dalapon
Dichloroacetic acid (DCAA)
Trichloroacetic acid (TCAA)
Bromochloroacetic acid (BCAA)
Bromodichloroacetic acid (BDCAA)
Dibromoacetic acid (DBAA)
Chlorodibromoacetic acid (CDBAA)
Tribromoacetic acid (TBAA)
MTBE
Mean
Recovery"
(%)
112
99.5
107
103
89.2
99.5
91.7
103
95.5
110
RSD
(%)
(n=8)
6.2
2.3
2.2
1.5
1.4
1.6
4.0
2.8
7.1
8.1
TAME
Mean
Recovery15
(%)
131
98.6
103
106
89.0
102
87.5
103
94.4
111
RSD
(%)
(n=8)
6.2
4.2
3.5
3.8
1.1
2.2
1.8
4.5
3.3
4.3
Recoveries were corrected for haloacetic acid (HAA) concentrations in the unfortified matrix, based on
an average value of three measurements. HAAs detected in the unfortified matrix included MBAA (0.14
|ig/L), DCAA (0.56 |ig/L), TCAA (0.32 |ig/L), BCAA (1.1 |ig/L), BDCAA (0.68 |ig/L), DBAA (2.2
|ig/L), CDBAA (0.91 |ig/L), and TBAA (1.0 |ig/L).

Recoveries were corrected for haloacetic acid (HAA) concentrations in the unfortified matrix, based on
an average value of three measurements. HAAs detected in the unfortified matrix included MBAA (0.22
|ig/L), DCAA (0.78 |ig/L), TCAA (0.36 |ig/L), BCAA (2.0 |ig/L), BDCAA (0.62 |ig/L), DBAA (4.0
|ig/L), CDBAA (0.82 |ig/L), and TBAA (0.79 |ig/L).
                                           552.3-41

-------
TABLE 9:   PRECISION AND ACCURACY IN CHLORINATED SURFACE WATER
             FORTIFIED AT 10 ug/L
Compound
Monochloroacetic acid (MCA A)
Monobromoacetic acid (MBAA)
Dalapon
Dichloroacetic acid (DCAA)
Trichloroacetic acid (TCAA)
Bromochloroacetic acid (BCAA)
Bromodichloroacetic acid (BDCAA)
Dibromoacetic acid (DBAA)
Chlorodibromoacetic acid (CDBAA)
Tribromoacetic acid (TBAA)
MTBE
Mean
Recovery"
(%)
97.8
98.8
99.6
98.0
99.8
101
103
102
105
108
RSD
(%)
(n=8)
3.1
2.0
0.90
1.1
0.80
1.4
4.3
1.4
5.5
6.3
TAME
Mean
Recovery15
(%)
97.1
98.8
97.7
97.6
100
101
104
101
105
106
RSD
(%)
(n=8)
3.8
2.0
0.42
0.42
1.4
0.93
2.3
0.93
3.2
3.4
Recoveries were corrected for haloacetic acid (HAA) concentrations in the unfortified matrix, based on
an average value of three measurements.  HAAs detected in the unfortified matrix included MBAA (0.14
|ig/L), DCAA (0.56 |ig/L), TCAA (0.32 |ig/L), BCAA (1.1 |ig/L), BDCAA (0.68 |ig/L), DBAA (2.2
|ig/L), CDBAA (0.91 |ig/L), and TBAA (1.0 |ig/L).

Recoveries were corrected for haloacetic acid (HAA) concentrations in the unfortified matrix, based on
an average value of three measurements.  HAAs detected in the unfortified matrix included MBAA (0.22
|ig/L), DCAA (0.78 |ig/L), TCAA (0.36 |ig/L), BCAA (2.0 |ig/L), BDCAA (0.62 |ig/L), DBAA (4.0
|ig/L), CDBAA (0.82 |ig/L), and TBAA (0.79 |ig/L).
                                           552.3-42

-------
TABLE 10:  PRECISION AND ACCURACY IN CHLORINATED GROUND WATER
             FORTIFIED AT 1.0 ug/L
Compound
Monochloroacetic acid (MCA A)
Monobromoacetic acid (MBAA)
Dalapon
Dichloroacetic acid (DCAA)
Trichloroacetic acid (TCAA)
Bromochloroacetic acid (BCAA)
Bromodichloroacetic acid (BDCAA)
Dibromoacetic acid (DBAA)
Chlorodibromoacetic acid (CDBAA)
Tribromoacetic acid (TBAA)
MTBE
Mean
Recovery8
(%)
126
113
112
102
92.2
106
105
111
103
99.2
RSD
(%)
(n=8)
4.5
3.8
1.1
3.7
1.2
3.8
6.1
5.3
8.8
7.3
TAME
Mean
Recovery15
(%)
124
98.9
87.3
107
103
104
106
105
103
103
RSD
(%)
(n=8)
9.5
2.6
2.7
3.2
2.1
2.4
4.0
1.7
4.5
3.2
Recoveries were corrected for haloacetic acid (HAA) concentrations in the unfortified matrix, based on
an average value of three measurements.  HAAs detected in the unfortified matrix included MBAA (0.07
|ig/L), Dalapon (0.19 |ig/L), DCAA (2.6 |ig/L), TCAA (1.1 |ig/L), BCAA (3.0 |ig/L), BDCAA (1.8
|ig/L), DBAA (3.3  |ig/L), CDBAA (1.8 |ig/L),  and TBAA (1.1 |ig/L).

Recoveries were corrected for haloacetic acid (HAA) concentrations in the unfortified matrix, based on
an average value of three measurements.  HAAs detected in the unfortified matrix included MBAA (0.28
|ig/L), Dalapon (0.35 |ig/L) DCAA (2.7 |ig/L), TCAA (1.6 |ig/L), BCAA (3.2 |ig/L), BDCAA (2.4
|ig/L), DBAA (2.5  |ig/L), CDBAA (1.7 |ig/L),  and TBAA (0.54 |ig/L).
                                           552.3-43

-------
TABLE 11:   PRECISION AND ACCURACY IN CHLORINATED GROUND WATER
              FORTIFIED AT 10 ug/L
Compound
Monochloroacetic acid (MCA A)
Monobromoacetic acid (MBAA)
Dalapon
Dichloroacetic acid (DCAA)
Trichloroacetic acid (TCAA)
Bromochloroacetic acid (BCAA)
Bromodichloroacetic acid (BDCAA)
Dibromoacetic acid (DBAA)
Chlorodibromoacetic acid (CDBAA)
Tribromoacetic acid (TBAA)
MTBE
Mean
Recovery"
(%)
104
100
97.8
95.9
98.2
100
106
101
110
116
RSD
(%)
(n=8)
1.8
1.4
1.1
1.0
0.81
0.53
1.8
0.89
2.8
3.4
TAME
Mean
Recovery15
(%)
99.7
98.7
97.9
97.1
101
99.8
106
101
106
105
RSD
(%)
(n=8)
1.7
1.1
0.63
0.83
1.4
1.4
2.3
1.6
2.8
2.6
Recoveries were corrected for haloacetic acid (HAA) concentrations in the unfortified matrix, based on
an average value of three measurements.  HAAs detected in the unfortified matrix included MBAA (0.07
|ig/L), Dalapon (0.19 |ig/L), DCAA (2.6 |ig/L), TCAA (1.1 |ig/L), BCAA (3.0 |ig/L), BDCAA (1.8
|ig/L), DBAA (3.3 |ig/L), CDBAA (1.8 |ig/L), and TBAA (1.1 |ig/L).

Recoveries were corrected for haloacetic acid (HAA) concentrations in the unfortified matrix, based on
an average value of three measurements.  HAAs detected in the unfortified matrix included MBAA (0.28
|ig/L), Dalapon (0.35 |ig/L) DCAA (2.7 |ig/L), TCAA (1.6 |ig/L), BCAA (3.2 |ig/L), BDCAA (2.4
|ig/L), DBAA (2.5 |ig/L), CDBAA (1.7 |ig/L), and TBAA (0.54 |ig/L).
                                            552.3-44

-------
TABLE 12:  SAMPLE HOLDING TIME DATA FOR SAMPLES FROM A CHLORINATED
             SURFACE WATER FORTIFIED WITH METHOD ANALYTES8
Compound
Monochloroacetic Acid (MCA A)
Monobromoacetic Acid (MBAA)
Dalapon
Dichloroacetic Acid (DCAA)
Trichloroacetic Acid (TCAA)
Bromochloroacetic Acid (BCAA)
Bromodichloroacetic Acid (BDCAA)
Dibromoacetic Acid (DBAA)
Chlorodibromoacetic Acid (CDBAA)
Tribromoacetic Acid (TBAA)
DayO
Rb
(%)
97.9
107
104
101
101
101
112
102
117
123
Day3
Rb
(%)
96.8
99.6
104
99.0
102
99.3
111
99.1
115
118
Day 7
Rb
(%)
98.5
105
103
102
98.8
100
113
100
119
124
Day 14
Rb
(%)
104
109
99.7
98.8
96.2
97.8
110
96.9
116
121
a All samples were fortified at 10 |ig/L. All samples were stored at 10 °C for 48 hours, and at 6 °C
thereafter.

bRecoveries were corrected for haloacetic acid (HAA) concentrations in the unfortified matrix, based on
an average value of three measurements. HAAs detected in the unfortified matrix included MBAA (0.30
|ig/L), DCAA (0.73 |ig/L), TCAA (0.33 |ig/L), BCAA (2.0 |ig/L), BDCAA (0.67 |ig/L), DBAA (4.3
|ig/L), CDBAA (0.91 |ig/L), and TBAA (1.1 |ig/L).
                                          552.3-45

-------
TABLE 13:  MTBE EXTRACT HOLDING TIME DATA FOR SAMPLES FROM A
             CHLORINATED SURFACE WATER FORTIFIED WITH METHOD
             ANALYTES3
Compound
Monochloroacetic Acid (MCA A)
Monobromoacetic Acid (MBAA)
Dalapon
Dichloroacetic Acid (DCAA)
Trichloroacetic Acid (TCAA)
Bromochloroacetic Acid (BCAA)
Bromodichloroacetic Acid (BDCAA)
Dibromoacetic Acid (DBAA)
Chlorodibromoacetic Acid (CDBAA)
Tribromoacetic Acid (TBAA)
Initial
R
(%)
99.0
99.7
99.0
97.9
99.3
100
99.3
101
99.7
102
Day 7
R
(%)
98.6
98.2
96.7
96.3
99.6
101
106
102
107
110
Day 14
R
(%)
100
98.1
98.1
97.5
98.5
100
102
101
102
102
Day 21
R
(%)
99.3
99.0
97.8
97.3
99.1
101
103
102
103
106
aExtract storage stability was conducted on three of the eight extracts from the precision and accuracy
study conducted in chlorinated surface water (fortified at 10 |ig/L) reported in Table 9. Sample storage
stability is expressed as a recovery value (%) calculated as described in Section 9.8.2. All values have
been corrected for background levels in the unfortified sample based on triplicate measurements as
described in Table 9.  Extracts were stored at < -10 °C.
                                          552.3-46

-------
TABLE 14:  TAME EXTRACT HOLDING TIME DATA FOR SAMPLES FROM A
             CHLORINATED SURFACE WATER FORTIFIED WITH METHOD
             ANALYTES3
Compound
Monochloroacetic Acid (MCA A)
Monobromoacetic Acid (MBAA)
Dalapon
Dichloroacetic Acid (DCAA)
Trichloroacetic Acid (TCAA)
Bromochloroacetic Acid (BCAA)
Bromodichloroacetic Acid (BDCAA)
Dibromoacetic Acid (DBAA)
Chlorodibromoacetic Acid (CDBAA)
Tribromoacetic Acid (TBAA)
Initial
R
(%)
98.2
98.9
100
97.5
99.6
101
103
100
103
104
Day?
R
(%)
95.8
98.7
101
96.4
101
102
104
101
106
108
Day 14
R
(%)
99.2
99.3
101
97.3
100
101
102
100
102
103
Day 21
R
(%)
98.3
98.8
100
97.4
102
102
108
102
110
110
Day 28
R
(%)
98.3
98.8
101
97.0
102
102
108
102
110
111
aExtract storage stability was conducted on three of the eight extracts from the precision and accuracy
study conducted in chlorinated surface water (fortified at 10 |ig/L) reported in Table 9. Sample storage
stability is expressed as a recovery value (%) calculated as described in Section 9.8.2.  All values have
been corrected for background levels in the unfortified sample based on triplicate measurements as
described in Table 9.  Extracts were  stored at < -10 °C.
                                           552.3-47

-------
TABLE 15:  INITIAL DEMONSTRATION OF CAPABILITY (IDC) REQUIREMENTS
  Method
  Reference
Requirement
Specification and
Frequency
Acceptance Criteria
  Section
  9.2.1
Initial
Demonstration of
Low Method
Background
Analyze LRB prior to any
other IDC steps
Demonstrate that LRB is
reasonably free of
contamination.
  Section
  9.2.2
Initial
Demonstration of
Accuracy
Analyze QCS

Analyze 4-7 replicate LFBs
fortified at midrange
concentration.
Recovery for analytes must be
± 30% of expected value.
Mean recovery must be within
± 20% of expected value.
  Section
  9.2.3
Initial
Demonstration of
Precision
Calculate standard deviation
and RSD for replicates used
in the Initial Demonstration
of Accuracy (Sect. 9.2.2)
RSD must be < 20%.
  Section
  9.2.4
Detection Limit
Determination
Over a period of three days,
prepare a minimum of 7
replicate LFBs fortified at a
concentration estimated to be
near the Detection Limit.
Analyze the replicates
through all steps of the
analysis. Calculate the
Detection Limit using the
equation in Section 9.2.4.
Note: Data from DL replicates
are not required to meet
method precision and accuracy
criteria. If the DL replicates
are fortified at a low enough
concentration, it is likely that
they will not meet precision
and accuracy criteria.
  Section
  9.2.5
Minimum
Reporting Limit
Estimate after determining
DL and prior to analysis of
Field Samples
At least 3 times the DL or at a
concentration which yields a
signal-to-noise ratio of at least
5.
                                           552.3-48

-------
TABLE 16:  QUALITY CONTROL REQUIREMENTS (SUMMARY)
  Method
  Reference
Requirement
Specification and
Frequency
Acceptance Criteria
  Section 8.3
Sample Shipment
and Storage
Samples must not exceed
10 °C within 48 hours after
collection.  Samples stored
in laboratory must not
exceed
6°C.
Sample results are valid only if
samples are properly collected,
preserved, and stored.
  Section 8.4
Sample Holding
Time
14 days with appropriate
preservation and storage
Sample results are valid only if
samples are extracted within
sample hold time.
  Section 8.4
Extract Holding
Time
21 days at < -10 °C
protected from light
(MTBE)
28 days at < -10 °C
protected from light
(TAME)
Sample results are valid only if
extracts are properly stored and
analyzed within extract hold
time.
  Section 9.3
Laboratory
Reagent Blank
(LRB)
With each extraction batch
of up to 20 samples
Demonstrate that all target
analytes in LRB are below V3
the MRL and that interferences
do not prevent the identification
and quantification of method
analytes. If targets exceed V3
the MRL, results for all
problem analytes in extraction
batch are invalid.
                                           552.3-49

-------
TABLE 16:  QUALITY CONTROL REQUIREMENTS (SUMMARY)
   Method
   Reference
Requirement
Specification and
Frequency
Acceptance Criteria
   Sections
   9.4 and
    10.3
Continuing
Calibration
Check (CCC)
Verify initial calibration
by analyzing a calibration
standard at the beginning
of each analysis batch,
after every 10 samples,
and after the last sample.

Low CCC - < MRL
Mid CCC  - near midpoint
in initial calibration curve
High CCC - near highest
calibration standard
1) The result for each analyte must be
70-130% of the expected value for all
but the lowest standard.  The lowest
standard must be 50-150% of the
expected value.
2) The peak area of internal standards
must be 50-150% of the  average peak
area calculated during the initial
calibration.
Results for analytes that  do not meet
IS criteria or are not bracketed by
acceptable CCCs are invalid.
   Section 9.6
Internal
Standard
(IS)
1,2,3 Trichloropropane is
added to the extraction
solvent.
Peak area counts for the IS in all
injections must be within 50-150% of
the average peak area calculated
during the initial calibration. If the IS
does not meet criteria, corresponding
target results are invalid.
   Section 9.7
Surrogate
Standard
(SUR)
 2-bromobutanoic acid
[80-58-0] is added to all
calibration  standards and
samples, including QC
samples.
Surrogate recovery must be 70-130%
of the expected value. If surrogate
fails this criterion, report all results for
sample as suspect/surrogate recovery.
   Section 9.8
Laboratory
Fortified
Sample
Matrix
(LFSM)
and
Laboratory
Fortified
Matrix
Duplicate
(LFSMD)
Analyze one LFSM per
analysis batch (20
samples or less) fortified
with method analytes at a
concentration greater than
the native concentration.
LFSMD should be used in
place of Field Duplicate if
frequency of detects for
target analytes is low.
Recoveries at mid and high levels not
within 70-130% or low-level
recoveries not within 50-150% of the
fortified amount may indicate a matrix
effect.
Target analyte RPDs for LFSMD not
within ± 30% at mid and high levels
of fortification and within ± 50% near
MRL may indicate a matrix effect.
                                           552.3-50

-------
Method
Reference
Requirement
Specification and
Frequency
Acceptance Criteria
Section 9.9
Field
Duplicates
(FD)
Extract and analyze at
least one FD with each
extraction batch (20
samples or less). A
LFSMD may be
substituted periodically
for a FD when the
frequency of detects for
target analytes is low.
Target analyte RPDs for FD should be
within ± 30% at mid and high level
concentrations and within ± 50% near
MRL.
Section
9.10
Quality
Control
Sample
(QCS)
Analyzed when new
Primary Dilution
Standards (PDS) are
prepared, during the IDC,
or quarterly.
Results must be 70-130% of the
expected values.
Section
10.2
Initial
Calibration
Use internal standard
calibration technique to
generate a calibration
curve.  Use at least 5
standard concentrations.
When each calibration standard is
calculated as an unknown using the
calibration curve, the result must be
70-130% of the expected value for all
except the lowest standard, which
must be 50-150% of the expected
value.
                                        552.3-51

-------
FIGURE 1:   CHROMATOGRAM OF THE HALO ACETIC ACIDS ON A DB-1701 COLUMN USING THE

            CHROMATOGRAPHIC CONDITIONS GIVEN IN TABLE 1 USING THE MTBE PROCEDURE
        Hz
      scoo-
      6IH10 -
      41X10 -
        0-
                              10
                                       12.5
                                                                              
-------
FIGURE 2:   CHROMATOGRAM OF THE HALO ACETIC ACIDS ON A DB-1701 COLUMN USING THE

            CHROMATOGRAPHIC CONDITIONS GIVEN IN TABLE 2 USING THE TAME PROCEDURE
       14000 -
       12000 -
       10JOU -
       4-nr.n -
       2000 -
         o -
         <
         <
         o
         Q
         ffl
                                                     
-------
FIGURE 3:   CHROMATOGRAM OF THE HALO ACETIC ACIDS ON A DB-5.625 COLUMN USING THE
            CHROMATOGRAPHIC CONDITIONS GIVEN IN TABLE 3 USING THE MTBE PROCEDURE
        Hz
       4om -
       10CO-
         o -
                                                                    CQ
                                                                    Q
                                               
-------
FIGURE 4:   CHROMATOGRAM OF THE HALO ACETIC ACIDS ON A DB-5.625 COLUMN USING THE

            CHROMATOGRAPHIC CONDITIONS GIVEN IN TABLE 4 USING THE TAME PROCEDURE
         11/'
        50UO -
        40110 -
        30110 -
        20110 -
                                 o
                                 [In
                                 O

                                 s
                                 o
                                 Pd
                                 CQ
a
H
                                           10        12




                                                    552.3-55
                                                               14
                                                                         16

-------