PB  198  065

CHEMILUMINESCENT  OZONE  METER FOR CONTINOUS
AIR MONITORING PROJECT

J0 B „  Tommerdahl,  etal

Research Triangle Institute
Research Triangle Park,  North Carolina

January 1971
                                        Distributed ... 'to foster, serve
                                           and promote the nation's
                                              economic development
                                                  and technological
                                                    advancement.'
    NATIONAL TECHNICAL INFORMATION SERVICE

-------
                                   FINAL  REPORT

                                       and

                                INSTRUCTION  MANUAL
                           CHEMILUMINESCENT OZONE  METER

                                        for

                         CONTINUOUS  AIR MONITORING PROJECT
                                 J.  B.  Tommerdahl
                                   R.  B.  Strong
                                  A. H.  Truckner
                                   January 1971
                                   Prepared for

                           AIR POLLUTION CONTROL OFFICE

                               Contract No. CPA  70-27
                  Engineering and Environmental Sciences  Division
                          Engineering  Physics  Department

-------
                            No.
                         APTD-0634
                                                   I
   Chemiluminescent Ozone Meter for Continuous Air Monitoring
                          Project
  Achcz(a)
    oB. Tommerdahl, R.B. Strong, A.M. TrucknerJ
           3. Recipient's Accession No.
                                                                      ง.' Repott Date
                                                                         January 1971
                                                                      8. Performing Organization Kept.
                                                                        No.
). Perfozsaiag Czeanisation Nome and Address              --  ซ       ..5     ,
   Engineering  and Environmental  Sciences Dfvt-sion-   uC-L-i j C/•  •
   Enginesri^g  Physics Department/JBesearch Triangle  Institute
   Research Triangle Park,  N.'C.    27709
                                                                      10. Project/Task/Wotk Unit No.
                                                                      11. Contract/Grant No.

                                                                           CPA 70-27
           Czganisetion Nome Qod Address
   EPAS Ar Pollution Control  Office
   Technical Center
   Research Triangle Park,  N.  C.   27709
                                                                      13. Type of Report & Period
                                                                         Covered
                                                                         Final
                                                                      14.
IS.
             Notes
 &. Abotroczo
  Ite&oreport describes  the results of a program to design,  construct,  calibrate and
  field test an Ozoner-specific instrument using a chemiluminescent principle.  Ttป*e  '">
  document  is also^ile  an instruction manual, with two separate attachments included;
  a set of  engineering drawings  for the machine shop fabricated parts and a subsystem
  manual comprised of  instruction  sheets and manuals for the individual commercially
  procured  units.   ;
 I, Key \7ci
-------
This report was furnished to the
Air Pollution Control Office in
fulfillment of Contract No.
CPA 70-27 by the Research Tri-

-------
          CHEMILUMINESCENT OZONE METER

                 Model EU-525-08
Engineering and Environmental Sciences Division
           Research Triangle Institute
             Post Office Box 12194

-------
                               FOREWORD




     This report describes the results of a program to design, construct




calibrate and field test an ozone-specific instrument using a chemilum-




inescent principle.  The work has been monitored by Mr. E. R. Lozano and




Mr. Frank McElroy, Air Pollution Control Office.




     The work was performed in the Engineering and Environmental




Sciences Division, Dr. R. M. Burger, Director.  Project Leader was




Mr. J. B. Tommerdahl.  Mr. A. H. Truckner fabricated and tested the




basic instrument.  Mr. S. R.  Stilley wired the amplifier, signal




conditioner and timing/control unit.  Mr. R.  B. Strong calibrated




the system and conducted detailed performance tests on the system,




amplifier and signal conditioner unit.  Mr.  C. E. Decker contributed




to the overall program in many areas.




     Some of the basic design of the instrument was carried out




under a program under contract with the Air Pollution Control Office




(Contract CPA-22-69-7).

-------
     INSTRUMENT IDENTIFICATION AND SPECIFICATIONS









Ozone Meter:  Model EU-525




              Serial No.   525-08
Photomultiplier Tube:




              Type - EMI-9558C




              Serial No.  20175




              HV          1090
                   Gain = 200 A/L




                   Dark Current =   0.8 nA   (Mfg. Rating)




                   Dark Current ?=   1.5 nA   (Measured)
HV Supply:   Power Design'Model 2K-10




             Serial No.   002066
Air Flow Setting:
69
12 vdc Supply:  Deltron Model  C12-2.8




                Serial No.     41605




+15 vdc Supply:  HP Model      60155C
                 Serial No.    961039

-------
                          TABLE OF CONTENTS

Section                                                          Page

  1.0   INTRODUCTION                                               1

  2.0   OZONE METER FUNCTIONAL AND OPERATIONAL CHARACTERISTICS     3

        2.1  Theory of Operation                                   3
        2.2  Functional Description                                3
        2.3  Physical Description                                  6
        2.4  Instrument Specifications                             8
        2.5  Front Panel Controls and Indicators                   9

  3.0   DETAILED SYSTEM DESCRIPTION                               15

        3.1  General                                              15
        3.2  Plumbing Subsystem                                   15
        3.3  Detector                                             19
        3.4  Calibration Unit                                     26
        3.5  Timing and Control Unit                              33
        3.6  Linear Amplifier                                     36
        3.7  Signal Conditioner                                   43

  APPENDICES

        A.    INSTALLATION AND OPERATION

        B.    MAINTENANCE, CALIBRATION AND ADJUSTMENT PROCEDURES

        C.    PARTS LIST

        D.    SUBSYSTEM MANUALS

        E.    ENGINEERING DRAWINGS

-------
                           LIST OF FIGURES

Figure                                                             Page

   1     OZONE METER FUNCTIONAL DIAGRAM                              4

   2     FRONT PANEL VIEW OF OZONE METER (WITH HV POWER SUPPLY        7
         COVER REMOVED)

   3     REAR PANEL VIEW OF OZONE METER                              10

   4     MAIN CHASSIS LAYOUT                                         11

         TOP AND REAR VIEW OF CHASSIS                                12

         LEFT AND RIGHT  SIZE VIEWS OF CHASSIS                         13
    (d)

   6     AC AND DC POWER DISTRIBUTION                                14

   7     PLUMBING SUBSYSTEM                                          16

   8     PLUMBING SUBSYSTEM PARTS LAYOUT                             17

   9     DETECTOR ASSEMBLY                                           20

  10     AIR FLOW PATTERN IN DISC CHAMBER                            21

  11     PM TUBE HOUSING WIRING DIAGRAM                              23

  12     INTERLOCK CIRCUIT                                           27

  13     CALIBRATION UNIT (TOP VIEW)                                 28

  14     OZONE CONCENTRATION VERSUS APERTURE SETTING                 29

  15     OZONE CONCENTRATION AS A FUNCTION OF LAMP CURRENT AND        31
         FLOW RATE

  16     CALIBRATION LAMP CURRENT VERSUS LAMP VOLTAGE                32

  17     CALIBRATION UNIT, IN SITU CALIBRATION                       34

  18     TIMING CONTROL  CIRCUIT                                      35

  19     PHOTOMULTIPLIER TUBE LINEAR AMPLIFIER                       37

  20     AMPLIFIER OUTPUT VOLTAGE VERSUS INPUT CURRENT FOR LINEAR    39
         RANGES

-------
Figure                                                          Page
  21    FREQUENCY RESPONSE OF AMPLIFIER FOR VARIOUS  GAIN         41
        SETTINGS

  22    AMPLIFIERS I0~7 AMP FULL SCALE, OUTPUT SENSITIVITY XI    42

  23    SIGNAL AND TIMING WAVEFORMS                              44

  24    BASIC SIGNAL CONDITIONING CIRCUIT                        45

  25    SIGNAL CONDITIONER CIRCUIT                               47

-------
1.0  INTRODUCTION

     The Research Triangle Institute has, under Contract CPA 70-27

with the Air Pollution Control Office, conducted a program to design,

conatruct, calibrate, and field test an o::one-specific instrument

using a chemiluminescent principle; the unit is intended, ultimately,

for CAMP stations.

     This document is a combined final report for the project and an

instruction manual for the ozone instrument, since a majority of the

information is pertinent to both.  Two separate attachements complete

the documentation; these are a set of engineering drawings for the

machine shop fabricated parts and a subsystem manual comprised of

instruction sheets and manuals for the individual commercially procured

units such as the high voltage supply and PM tube housing.  Thus this

is, in general, a comprehensive engineering description of the design,

construction details, calibration techniques and operational charac-

teristics; of the ozone monitor.  Material essential in an instruction

manual has been expanded, particularly with respect to the design,

testing, and operational characteristics.

     The ozone meter developed operates on the chemiluminescent prin-

ciple.   When ozone reacts with certian organic compounds, the reaction
                                           o
produces a minute luminescence in the 5800 A region which can be

measured by means of a sensitive photomultiplier tube and dc amplifier.

The chemiluminesence principle for ozone measurement has the advantage

of being essentially specific for ozone for the expected concentrations

of other pollutants.

     The ozone meter was designed as a laboratory instrument for the


-------
the meter is considerably more sophisticated in design than would




be necessary for a meter that is designed as a production unit.




This versatility in design provides for a wide latitude of test




conditions.




     A functional description of the ozone monitor is given in




Section 2.0; and a detailed description of the design for the




various sub-assemblies are presented in Section 3.0.  Various




details concerning parts list, installation and maintenance of




the instrument, calibration procedures, and operation procedures





-------
2.0  OZONE METER FUNCTIONAL AND OPERATIONAL CHARACTERISTICS






     2.1  Theory of Operation




          The principle on which the chemiluminescent ozone meter




operates is as follows:  when ozone reacts with certain organic




compoundss a minute amount of light is emitted.  The amount of light




is directly proportional to the ozone concentration, and the wave-




length of the emitted light is a function of the organic compound




or dye used in the formation of the chemiluminescent disc.  Gases




to be sampled are passed over a disc coated with the dye.  The emitted




light resulting from the reaction with ozone is detected by means of




a sensitive photo-detector.






     2-2  Functional Description




          A functional diagram of the ozone meter is shown in Figure 1.




The basic element of the system is the chemiluminescent disc and




photo-detector assembly.  The gases to be sampled are metered at a




constant flow rate across the disc.  The emitted light resulting from




the reaction of ozone with the chemiluminescent material is monitored




by the photomultiplier tube.




     The system has essentially three modes of operation.  During the




MEASURE mode gases are pulled through the sample-gas inlet, and then




passed over the disc.  During the CALIBRATE mode the sample-gas inlet




is closed and the calibrate line is opened by means of electrically




actuated solenoid valves.  Air which enters the calibrate line passes




through a filter which removes contaminants and destroys any ozone





-------
   LAMP
  CONTROL
MODE SW j

JHttt
115 V - • >- TTMTWP •

i
•J H
HO;
• a 2 V
M , ฐ
U W _5 i
J 0 j
O CO ;
CAL AIR 	 ' * SHUTTER APEF
INLET ' "' A CONTROL-^ AT


—~ 2UU ml/min j
^ UNIT
SAMPLE AIR
INLET
^-. 11C TT CONSTANT RANGE
— vi/ - •L-L-) -/ _,. ^T ^
115 V 115 v ' ! I
LAMP ^ 1 1 U
UPPLY ~ *" POWER • *•'< AMPLIFIER 1
SUPPLY ! , 1"

-*• SIGNAL
i COMDITIONISG
i * ' 1 Q
i \ \-
** ! HV i ' V
[ETER . i... ..._V 	 ; | 1
•^_L, , SIGNAL PANEL ( / }
"-T" 1 i i
ru,
^TURE ; w 1
)J --. 1 LAMP ! g
-> I s SHUTTER FLOW
''^ • -!-ฃ!__, CONTROL --> METER -? ,
'^_ '-~ — ; ;— H -*"" n^ TV
! :. i ; i ( i i \ y"
1011 l ' DISC ' IFILTERB! 	 1 :— V\,

i HOUSING ; - ->.'" 	 "' 	 i i
1 200 ml /min :
- 1
COARSE -
FLOW
ADJUST
iETER <^>
~ OUTPUT
FINE
FLOW
\DJUST 115 v
i *
tt i PUMP !
fl
• EXHAUST

-------
as it passes through the calibration unit by exposing the air stream
  \
to an ultra-violet light source.  A calibrated aperture control

facilitates varying the ozone concentration.  A front panel control

and meter provides for lamp current control and monitoring, respectively.

During the PURGE mode, the unit operates as in the CALIBRATE mode,

except the calibrate lamp shutter remains closed.

     The gases are metered at a constant flow rate of 200 ml/min

through the disk chamber  by means of control valves, a flow meter

and pump.  The flow meter and vernier control valves are located

on the front panel of the instrument.  The timing and control unit

provides the necessary switching voltages for actuating the various

solenoids.  Selection of operating modes is made by means of front-

panel push-button switches.

     The light emission from the disk is detected by means of a

sensitive photomultiplier tube.  The output of the PM tube is

amplified and presented on a panel meter and at a recorder output

connector.  Five linear-decade current ranges are provided.  Contin-

uously variable controls are provided for span and zero adjustment.

A time-constant control is incorporated in the amplifier which

facilitates smoothing of the analog data.  A manually operated shutter

incorporated in the disk chamber allows for monitoring of the PM  tube

dark current.  High voltage for the PM tube is provided by a well

regulated, adjustable HV supply.

     The output signal-to-dark current ratio for a given concentration

of ozone is a direct function of the sensitivity of the disk and  the


-------
of the emitted light.  Since the dark current and the quantum efficiency



vary considerably from tube to tube  one may expect a factor of 10 or



more in this ratio.  With tube selection this may be reduced to a 10-20%



variation from unit to unit.



     The system design includes two selectable sampling modes; SAMPLE-I



is continuous sampling while SAMPLE-II inserts a PURGE mode for a 30



min period of time, every 12 hours.  The sampling-mode consists of a



75 sec-measure, 75 sec-purge, 75 see-calibrate, and 75 sec-purge cycle.



In addition, push-button actuated modes are available'for operating



continuously in the RUN, CALIBRATE, or PURGE modes; these facilitate



test and calibration procedures.  An EXTERNAL mode is provided for



remotely controlling the operational mode of the instrument.



     The output signal from the amplifier is provided at an output



connector and at the input to the signal conditioner.  This latter



unit accepts the sampled signal and essentially provides an output


signal, e , which is


                                   m-p
                              Q  =: 	*—
                               o   c-p


where     m - is the measure signal value



          c - is the calibrate signal value, and



          p - is the purge signal value.



     2.3  Physical Description



          A front panel view of the ozone meter illustrating the type


and location of all controls,  meters,  etc.,  is shown in Figure 2.   The



cover for the HV supply,  which is normally in place,  is shown removed
                                      r


-------

-------
     There are two basic units, the main chassis and the HV power

supply.  The latter is located in the bottom of the cabinet.  The

main chassis is mounted on chassis slides and may be readily pulled

out on the chassis slides for adjustment and testing purposes.  A

rear view of the instrument is shown in Figure 3.  A plan view of

the main chassis is given in Fig. 4, showing the relative positioning

of the major components and assemblies.  Top and rear photographs

of the chassis are shown in Fig. 5(a) and (b), respectively.  Side

views of the chassis are shown in the photographs in Fig. 5(c) and (d).

     Both the main chassis and HV power supply fit into the standard

19" width cabinet.  The overall physical dimensions of the instrument

are as follows:

                           Height - 47 cm,

                           Width - 50 cm,

                           Depth - 58.5 cm,

                           Weight - 68 kg.

     The unit requires approximately 300 watts of power from a 115 V,

60 Hz single phase source.   A block diagram illustrating the AC and

DC power distribution is given in Figure 6.


     2.4  Instrument Specifications

SYSTEM:

     Dynamic Range - 0.001-2.0 ppm
     Resolution - 1 ppb
     Minimum Detectable Level - 1 ppb
     Accuracy - + 10% at time of calibration,  with respect  to values
                obtained by KI technique (See Section  B-ll)
     Stability of calibration signal - approx.  1.5%/volt of  line  voltage
     Calibration - nominal  range 0.1-0.2 ppm
                   internally adjustable 0-0.1 ppm
     Output - front panel meter 0-10 units

-------
 AMPLIFIER;

      Output Meter - full scale value indicated by SENSITIVITY and
                     FS OUTPUT control settings
      Sensitivity  - linear - 10~9 amps, FS

                              10"8  "

                              10"7  "
                              10-6  „

                              10~5  "     "
                                                               -9
      Stability    - linear - 0.01% perฐC; 1% per month (for 10   amp range)


      2.5  Front Panel Controls and Indicators

 MODE CONTROLS;  (Lamps indicate mode of operation)

      Ext. - external timing control
      Cal. Cont. - continuous CALIBRATE
      Meas. Cont. - continuous RUN
      SAMP. I - five minute cycle;  75 sec measure, 75 sec purge,
                75 sec calibrate, 75 sec purge
      SAMP. II - same as Samp I, except a 30 min purge mode is inserted
                 once every 12 hours

 AMPLIFIER:

      Span - modifies full scale sensitivity by factor indicated
      Time Constant - normal, 1 sec, 2 sec, 4 sec             R
      Zero - provides dark current offset control from 0-10   amps

 LAMP CURRENT - controls calibration lamp current

 ELAPSED TIME METER - hours and 1/10 hours, actuated during power ON
                      condition

 FLOW CONTROL - valve and flowmeter for flow control and monitoring

 SHUTTER - (Lamps indicate shutter position) open for operation; close
           for dark current check and disk removal/replacement

 DISK ACCESS DOOR - allows removal/replacement of disk

 INTERLOCK SYSTEM - indicates shutter control status; warning buzzer when
                    disk access door opened without fully closing shutter

 HV POWER SUPPLY;

      Power Switch - ON position
      HV Controls - normally set for value which yield PM tube gain = 200 A/L
      Cover - normally left in place

AC POWER - single switch turns entire unit on

                                  \ 9
                                 •  I

-------
                            REAR ACCESS  DOOR
                                                                      FAN
                                                                                    10:'.V
                                                                                          1
             PUMP
            BYPASS
                                       O
              o     o
   SIG COND
     OUTPUT  OUTPUT
SAMPLE
INLET
EVENT
MARKER
                                      AC POWER
                                       CABLE
i
L	:
                         FIGURE 3,  REAR PANEL VIEW OF OZONE METER

-------
    SAFETY CHAIN ATTACHMENT POINT
         + 12 VDC
       POWER SUPPLY
     (+15 Vdc Supply)
j (SIGNAL
  COND.
  UNIT)
     UV
    LAMP
    XFMR
  I

 A'C
PdwER
STiRIP
  I
                                      REAR
                       j
              TIMER AND CONTROL
                UNIT ASSEMBLY
                                      SAMPLE
                                      SOL
                                      VALVE
                               CAL
                               SOL
                             VALVE
                                                             PUMP
1
                                                         CALIBRATION UNIT
                            AMPLIFIER
                            ASSEMBLY
                                                r
                                                       DETECTOR
                                                       ASSEMBLY
                                      FRONT

         (     )  indicates unit in dotted lines which are on a 2nd level



                         FIGURE  4.  MAIN CHASSIS  LAYOUT


-------
ซ.ป.

             (a)
(b)

-------
                      Cc)

                                                  \
                      (d)

-------
 117 VAC
 60 Hz
                  5 AMP
r
LEGEND
   [)  AC PLUG
  -CO- IN-LINE PLUG
       BNC
       DIRECT WIRING

1
1
1
1
1
1
1

1
1
I
j
1
1
1
1
1
1
1
1
1
1
1
1
1
1










L_


i i
-ID-1
_. m 	

1
i
1 AUTC 	 j,
1 XFMF
1
1
Irv
| J
\ >
1
'Pi
u
1
1
,D
i
p\
[)



i
(
r\
i r
LS
1
1 ป-
1
*\
r^
\s

•to THTMD


^
\ tAN

CALIBRATION
T AMD
LAMr
SUPPLY



IN POWER
SUPPLY

RELAY
POWER SUPPLY
12 VDC
' r

TIMING
rONTROT





AMPLIFIER
POWER SUPPLY


SIGNAL
CONDITIONING
POWER SUPPLY






-25 mA
* /u v m \JL\ >- uv LAMP
60 Hz ^^



VUU. . PTinTOMlTT TTPT TFU T



> SOLENOID VALVES
.. . ,. ... ._ v PANFT T TPHTI

1 9 \TTปC



13 5 VAC
^ INlKKJ-iUCK. CIRCUlJ


+15 VDC
*" AMPLIFIER.


+ 15 VDC
* SIGNAL COND. UNIT

-------
3.0  DETAILED SYSTEM DESCRIPTION






     3.1  General




          A functional description oi: the overall system was given in




Section 2.0, reference should be made to this section for clarification




of the individual subsystem functional requirements.  Throughout the




system,, commercially available parts are used where practical.  In




some cases it was necessary to modify slightly some of the components.




All call-out parts may be identified by manufacturer and type by




referring to the respective section in the parts list (see appendix).






     3.2  Plumbing Subsystem




          A detailed diagram of the plumbing subsystem is shown in




Figure 7.  In this diagram the functional location of all significant




components are shown, along with the type of tubing, fitting, etc.




A parts layout for the plumbing subsystem is shown in Fig. 8 in




which the relative location of all the components are shown.




     In the MEAS mode, gases are pulled through the sample air inlet




line; during the CALIBRATE mode, gases are pulled through the calibrate




air inlet line and through the filter.  This latter operation removes




contaminants from the intake air and destroys all of the ozone.  Both




the sample and calibrate gases follow common paths after passing point




(7A).  Only one of the two ports is open at any given time.  After




passing point (8) the incoming gas (200 ml/min) is passed through the




light rejection coil into the detector unit.  The light rejection




coil is a single turn of teflon tubing—covered with black-shrinkable

-------
     p
\     INLET
CALIBRATION


UNIT


                                                                                                REJEC1
                                                                                                 COIl
                      LEGEND
                           NEOPRENE
                           COPPER

                           TYGON
                           GLASS
               ZZ.~Z.rH~  QUARTZ
                  o
 PARTS CALLOUT
(Stt section 5-3)

-------
          Pfcj
          TUBE
        HOUSING
 "LIGHT
 REJECTION
   COIL
                                               FILTER
                                                                              COARSE
G   SUBSYSTEM

-------
                                      (FRONT PANEL)
                                                                    PUMP
    /
  LIGHT
REJECTION
  COIL
                   FIGURE 8.   PLUMBING SUBSYSTEM PARTS LAYOUT

-------
vinyl tubing.  This device reduces to an acceptable level the amount




of external light entering through the entrance port.  The gas passes




over the chemiluminescent disk, through another filter, flowmeter,




valves and pump.  A bleed valve (coarse control) is provided to




prevent the pump from being overloaded.




     A flowmeter is used for adjustment and monitoring; the meter has




a standard accuracy of + 10% of maximuir. scale from 100% to 10% of scale




reading.  The calibration curve is shown in Appendix B; see this section




also for instructions for proper set-up procedure.  Two precision




vernier control valves are used for this purpose.  The fine flow control




valve is located on the front panel of the instrument.




     All parts of the system between the sample inlet port and the




detector are constructed of either glass or teflon in order to prevent




the destruction of ozone.  Soft copper and tygon tubing, brass tubing




fittings and brass valves are used where contamination is not a




problem.




     Two filters are used in the chemiluminescent ozone meter; these




are shown functionally in Figure 7.  Filter A is a disposable milli-




pore aerosol filter which has been modified and includes filter paper,




glass wool, activated charcoal, and manganese dioxide.  Its function




is to destroy and prevent ozone contained in the ambient air stream




from entering the ozone generator.  This is required to insure o




constant part per. million calibration ozone output from the generator.




Filter B (Figure 7)  is a disposable millipore aerosol filter filled




with glass wool.  Its only function is to prevent particulate matter

-------
from entering the rotameter and needle valve.  A cutaway drawing




of Filter A and instructions for care and maintenance of both filters




are given in Appendix B.




     3.3  Detector




          A cross sectional view of the detector assembly showing the




principal components is shown in 'Figure 9.  The location of the




detector unit in the system is shown functionally in Fig. 1, and




the associated plumbing is shown in Figure 8.  The principal components




of the detector are the photomultiplier tube (PM) and HV supply, disk




chamber and associated plumbing.  The chamber is designed so that the




inlet air enters the chamber tangentially to the disk and follows




a spiral-like flow over the disk, passing through the exhaust port




which is located in the center.of the disk chamber.  A pictoral




representation of this is given in Figure 10.  The top of the chamber




consists of a quartz window sealed to the chamber, thus providing an




air-tight chamber.




     Light entering the chamber by the inlet port is reduced to an




acceptable value by means of a loop in the inlet glass tubing.  The




loop is encased in black vinyl heat-shrinkable tubing.  Black rubber




tubing satisfies the requirements for the exit port.




     The cutaway view of the PM tube and housing in Fig. 9 shows the




placement of the special grounding braid located between the tube




near the cathode end and the MU-metal shield.  The latter shields




against stray electric and magnetic fields.  The braid reduces noise




due to static charge buildup on the tube.  This is an addition to the

-------
PM TUBJ

PM TUBE
HOUSING
MU-METAL SHIELD

GROUNDING BRAID

INSULATOR
-<"
WINDOW
       REMOVABLE PLATE
        (DISC HOLDER)
                        EXHAUST
                         PORT
                     (TOP VIEW)
           FIGURE 9.  DETECTOR ASSEMBLY
                                                          GROUNDING
                                                            BRAID

                                                          MU-METAL
                                                          "SHIELD
                                                         MOUNTING
                                                         i PLATE
                                                         PLATE
                                                         "OPENING

                                                         DISC
                                                         PM TUBE

-------
FIGURE 10. AIR FLOW PATTERN IN DISC CHAMBER

-------
 standard PM  tube housing.  The separation distance between  the PM  tube



 cathode and  the chemiluminescent disk  is approximately 1/2  inch.   See



 Appendix B.10  for details on  tube handling.



     The cathode of  the PM tube is positioned approximately flush  with



 the  face of  the PM tube housing bolt flange.  A circuit diagram of the



 PM tube base wiring  is shown  in Figure 11.  The tube is an  EMI Type


                                   -9
 9558C  selected for dark current <10   amps.  The high voltage (HV) for



 the  PM tube  is generally set  for the value which yields a gain = 200 A/L.



     A complilation  of the PM tube characteristics for several 9558C



 type PM tubes  is given in Table 1.  This includes the factory speci-



 fications and  the measured values for dark current.  The ratio of



 measured to  rated dark current was approximately 3/1 which was in



 line with the  expected increase in PM tube dark current due to



 increased temperature.  The factory rated value is run at 20ฐC while



 the  units in the laboratory were operated somewhat above this level.



     The signal-to-dark current ratio, which was used as a measure



 of comparative sensitivity for the PM tubes, was measured on the



 same unit, same disk, etc., with no changes in conditions within



 practical limits.



     The high voltage supply  is a Power Designs Model 2K-10 with front



 panel  controls which allow for precise selection of the high voltage



value.  A cover is provided for this unit so that the voltage controls



will not inadvertently be changed.   The unit is so designed so that



both the high voltage and primary power ON switches are left in the



ON positions and the power for the entire instrument controlled

-------
NJ
U>
                              MU-METAL
                               SHIELD
              HV CONNECTOR
             - RI;L - 200K, 1% METAL FILM
          Dl - ZENER DIODE   1N5276B

                 Vz  = 150 V


                 V_ = 0.85 mA
T&
                                                                                     ft,,
                                                                                                     p 12.
                                                                                                      •-.  OUTPUT
                                                                                                      )   BNC

-------
TUBE 9558C DARK
SERIAL # CURRENT
(RATED)
11781
11959
11987
11814
11994
11312
20002
20175
2.
0.
1.
3.
0.
2.
2.
0.
6
8
7
0
5
2
8
8
DARK RATIO OF
CURRENT QUOTED TO
(MEASURED) MEASURED
6.8
2.4
4.7
10.0

6.2
0.5
1.5
2
3
2
3

3
0
1
.62
.00
.76
.33

.55
.178
.88
HIGH CATHODE
VOLTAGE (FOR SENS.
GAIN OF 200) yA/L
1930
990
1130
1970
1030
1340
1110
1090
180
94
118
170
110
96
98
111
SIGNAL TO COLOR
DARK CURRENT SENSITIVITY
RATIO B R IR
54.0
100.0
16.0
23.0
180-200
16.7
128.0
	
10.3

8.2
9.0
7.4
7.0
7.1
8.6
52
26
41
55
36
34
30
30
4.2
0.4
1.6
5.8
1.2
2.6
0.6
0.7
Unit used in Model EU-525-08

-------
by the one power switch.  A test was made on 5 HV supplies to




determine the accuracy of the voltage setting controls.  This was




done by setting the controls for the 500, 750 and 1000 volt levels




in turn and measuring the true output with a Fluke Model 895A DC




null voltmeter.  The results are shown in Table 2.









                               Table 2




                    HV POWER SUPPLY SETABILITY TEST

Unit Number
9C2151
902166
902185
902193
902163

500
501.4
501.6
501.94
501.65
501.43
Vcltage Setting
750
751.5
751.9
752.5
751.94
751.71

1000
1001.8
1002.2
1002.985
1002.28
1001.96
     A shutter is incorporated in the disk housing between the quartz




window and the cathode of the PM tube.  The primary purpose of the




shutter is to facilitate the measurement of the PM tube dark current.




In addition it is used to protect the tube from external light




when the disk is replaced.  Access to the disk is through the door




on the front panel.  A miniature lab jack is used to raise and




lower the disk holder.  An interlock circuit, designed to prevent




inadvertent exposure of the PM tube to excess light, gives visual




indication for the fully open and fully closed conditions for the




shutter.  When the shutter is in the fully opened position, a mechan-




ical stop prevents opening of the disk access door.  A buzzer sounds

-------
if the disk access door is opened when the shutter is in any but the

fully closed condition.  Details on this circuit are given in Figure 12.


     3.4  Calibration Unit

          In the calibration unit clean air is exposed to a constant

amount of ultraviolet radiation from a low-pressure mercury arc lamp.
                                                          o
The particular spectral region of interest is around 1850 A.  The
                                  o
resonance line of mercury is 2537 A; however, enough energy is radiated

in the shorter wavelength region to produce the desired ozone level.

The calibration unit consists of a low pressure mercury arc lamp, a

quartz tube through which the clean air flows, and an adjustable

aperature and solenoid controlled shutter.  A cutaway drawing of the

calibration unit listing the major components, is shown in Figure 13.

Details on lamp replacement and aperture range adjustments are given

in Appendix B.5 and B.9, respectively.  The level of radiance is

controlled and monitored by means of a variac and lamp current meter.

Various, but repeatable, levels of 0- concentration are obtainable

with the adjustable aperture.  The specific calibration curve for

the instrument is shown in Figure 14.

     The housing is essentially a sealed unit to preclude the

escape of ozone or UV radiation.  Because of the thick wall con-

struction, the unit is reasonably stable, thermally.  The lamp is

left on at all times and the unit is effectively turned on and off

by activation of the electrically operated shutter.  The particular

UV lamp used is rated at 17 mA.   The lamps are pre-aged for 400-500

hours and then selected for optimum characteristics.

-------
                           1
.**•
.^-
i""
V|
i i
i i
TB I ! ONO
	 	 - A ^ ' S901 ' — <""
f — - "-XI us\JJ- ^
i RED X
	 	 i H NC
110 VAC , ( r.^ 12>6 VAC i pu^ 0pฃN
60 Hz -- | ;=ป ฉ LAMP
1 ^ Bi
	 	 i ..-, i
TB.
4
. 	 ..,<&.. . — , i 	
SHUTTER CONTROL ROD
I
j ' " 	 1 DISC ACCESS
: j 1 ^ DOOR SWITCH
ฃ NO ; 1 ^ NO
1 S902 L^r>'
S903
0 NC •; NC
i
i FULL CLOSED 4 TB5
4 LAMP .isi-
: B2 [3 tf-i BUZZER
i TR iji.ii.jJ
1 TB4 f
   GREEN
NOTE:  SWITCHES SHOWN WITH SHUTTER OPEN AND DOOR CLOSED

-------
       HOUSING,  IRRIDITED
         2024 ALUMINUM
ALUMINUM
REFLECTOR
ROTARY SOLENOID
    SHUTTER
  1.75"
  VARIABLE
DIA APERTURE
                                                                          PEN RAY
                                                                          UV LAMP
                                    MOUNTING PLATE
                                    &  LIGHT  BAFFLE
                                   QUARTZ TUBE
                                     10 mm OD
                       FIGURE 13,   CALIBRATION  UNIT  (top  view)

-------
::::::::::::::::::ti::::;::::s:i:::
                          tซHซKlH(ปJiซ,JWlt=ซ*ปปrซซปซปB"?->ซปP"ซปซซKlilซป*-Sซ..ปซปB'l^*lป'*S.f5mB-'5 "^ If ••ป '...•.ซ.* i ป-•  ,   '--...  .  .. .     ,., ••..
                          ซ*tf •ซ••••••***•***•ซ*• **WซV**ซ0*ซ ซฃ*•ซ•ซ*•* 4ffttfAv  3 ^ ^- <; 9 "   i^^^t'* :,ป**•" i1 ••* ป , -* --* ป •  : .  ...      • •- -
                                                                                         _                          '- " * a * ซ * •' r i' -•-: 1- a ซ ป• -- '* * a •> * -.' ^ * J- - '.• * ป ' • s ป n • ป ft * - r I- * •> . . .* ->i>. *ซ•••ซ>••**
                                                                                        ปป*:*i"HH^HIllปIHkf*ป*"IT ป'ซ••• ' "ป'ป.'- ป!!-.ซ>. ป •(. B.ซ s ••>ป>. FV * ปa ซ a* * K ..ป*ป*, .ป r. r * I ..ป t i. .='.,.*,. .r.k,..*^...j.*..
                                                                                                    -     .-   .  -         .         •          -.            .
*• ซ av ••fl •* •*• V *wซG^B !•••ปป• n ••*••>• MB* ^•••••ii'ซ*e*ซr'*a** ป AP0M'Jป'JIA>B
	KABBM*ซ•ซ•*•••• {**••ซ••••ซ•*•••••ซ* •••*•••••••••••ซ•*••••ซ•
                                                                                                .   ^-     ,
                                                                                   •55"pปปซS!IJ"ซCซ*S^ปr ,-*^-**~, -:, '.i...  '- !,.:" *.*ซ$.,*, f .,„[• i* ..
I          **ซ••••••!ซ•••ซ••••••ซ
          •••••••nil••••*••*ป•••
          ::::::::ป::::::::::::
          • •*•ซ•ซ*•  ป•••ป•*ปปป•!***'•ซปป** *ป••ปซ•%ซ.<•>•• ft ปSa.-Bf1 !'* ^S^R *ป'1 tป *ซซป•ซ ป•"ป! ,.,,.,ป.ปปป ****ป*ซ*-* ป,,ป...ป,*ป, •- -.,.,, •ปป-.il*4-,lH... ซ••.,ป•.„..-••ซ-,.•• 'V .. -.-.M-il~ •-.-..-•,.I .. ..,!'ป..-.,.. , !>..ซ.'
          KBM*ซซK<*ซ  • •••••••.•tt**lilซfta**KllfcKB*llli*ซ-' - *S ••* * J? si •• * *  • r • <• • e ' e  •*••!ซ••{•ป 4 ••••• ซk ••ซ••••••••••••* f *•  >Hป*ซ4  ".••.I *ป.-...     • .    .     .. ,..ซ,<*ป*
          • •••ซ••••  !•••••••••••!••••••• •••••••••B ••*••••*•*ซ••••••••• ••ป•!ซซซ•ซ• 4 >**•ซ*•-ซ t M  •.  '  . I. <-  '     .         , * *•ป ^.iBI-.^..,!      .     ..mil...  -••'••'...   •  .,, . r     .,----„,ปปป
          :::::::ป  ;::::::  :::::  :    :::::::::::::::::::::::::::::::::::::: :::  ::::;••::::::::%:::::::::::::::::::::::ป::;::::::    •--.-. ••   ::::E:::u!:::::::::::::
          :::::::::::::  :  a:::  ;    ::j::; :  ::           :::: ::s::: ::!::::::::::::::;:::;::• ;:::::::::::;:;:::::::.".::::::.":::::   ;:;:•:    ;.;  i-:;;::.-;;::.-.':::;:
          • ••••••••  KMtt J1  ul  Bซป ปB  ป    ป•!*ปป ซ  ^ป           ซ-•ป ปW *ซปfl," * ***=*, ป ป "T. !„>-->'•   •..(.•..ปป,•  .-..Tซt.,,i,.,..i^iป.ซi,ป|.ป. •-ซ,•--ป-  .,..-..-   *ป.. wn    *.  -.•*••••••*••*•••*•
          :::::::ป  ::!:  :  H; ::  :    :;!::::::           :: n s:;,r •                    .  ::::\::::::h:::::::::M::::::::::::::::::-    as a         ::::::::::::::::::
          ::::::::i  ::::  :  :::::  :    :::::::::           :::: ป::J                  ..::::::::.:::::::;::::::::::::::::                       ::          ;::::::::::::::::
          ฃ••••••••  (•• •  B  •••••  •    *ปซป$BW*6           ซea>ปซปXMl  	i.: ...-1' .' ." •.•-„.= ซ!,••.ซป."  i- •.,'.--    --   .'•    •.-ซซซ    tป  ซ ^ K jm.ป^.. -(.i.f.'  , -
          • ••lป^P"il  *ซป ป  W  ปiiป ซt>  ป    1*MซปปซI .>  ซป           *ป #9 ., *K. •, *•*•&  *•• ป  •  •ฃ• BB  •    •••••• •  ••           •• BB ^PMปBM •   E> ซ * wซ |.i>,a,ป.ซ..> ••illBBซi&Mi>Hซป>**kปB*>**"* ป~* ** I-..:-x 1 .•••ซซ=•ป--- I-:-'  ซ , ปป* ป I. K* * ^ •-;. -  .-  .  .•  >
           • • , ll*ซCS                                     !••!. •
                                                 !ป**•

        '•* •••••••••I *•*•ป••ปซ ซปปป•*ปปป* •ป*ป•ป• ii •*ปป
        -• •'••••>• ••••••••••••••••'^•••vn an *••••••••'ซ••
        .•••' •• *ป tซปJ •••••••••ปปปซKซซปปป*ปซปlซซf*f **••*
        !••••••• •••••••••••••••••ซปJ*llปPI>P*
        !•ป•••••••••••ป*••••••••••••ป••••••*••••••••**
        '•••ซ••ป•••**•ซซซ*ซป••ซ*•ป•ป**ซปป••*ป•ป**ป•••ป
        i::::::::!::::::::::::::::::::::::::: H::i::


        ••••ป••••••••••••• •••••••••*•••-••••• •••••••
        • •••<-ซ* ซ••••••••••••••••••ซ••***••*• •••!•• flit
        ::";:;','ป:::::::::::::::::::::::::: :::::::
        •••••••*•••*•ซ•••ป                         ซ•
        *•••• •••••• •••••••^•^••••••r •••••••• V*B* JVB
        ป•ซ•• •ป••ป* ป#ป•ป*           .*•<•••** •*• -•ป•
                                                 •
                                                                                                                     .,,.                      ,,            .
                                                                                             "" w*J,'ป ป *- "H^'B * ,ป *L- "*' i'" '  '''"';   ' .    . , '*' ..,-',„„.,'" * „ "-',ซซ.. JJi, *L^!i* i  •-ซ!-• ^i:*.; M;^
                                                                                              ...,,,,;,  ........v.,....,, ..„....,..•,. ;;,.„,,,,.,,..,,.„..,.:,;;.,:.,,:•,, ,;::
                                                                                             *!"*.-* -i -1 — '*•.>!•• •••••••••••••I


                                                                                             • ,    I-      ,i.  ,  1^1    .  ,"•  ,. T i- * 1 I. *


                                                                                                   -   '                   '
                                                                                                 i ,      •     '  "i -,-.•• mป*j •,ป:•ป-
                                                                                                 ,-  .  •   .    --  -    „ ,.ซ* .  .,ป .
          ::::::::::::::::
                            BซJl* ซ••••••••ป•••••••*•<•••*•••
                            •• ป *ปป•••••• •••••••••• ปปป•••••
                            ""*"""""-'"" ***!*ES!:::";^-;i;:
                            BB ••••••!!••*••••• ••••••••••••
                            ••! ••••WHBPBBBPBIJI •:•••• jl*BVBilB


                              il!iiHy!i!n!i:ili
                            ::::::":::::::::: ::::::::::::
                            ::::::::::::::::: •ซ'•""*
:	-•.•••••ซ••••ซ **-.N. ซ>•*ป* ซBปl.i|liXป*ปo. I. Jlป ป= ..ซ.!. i' *ป.1. e ซ••••••••>*••••••••••ซ ••ป•ซ-----
:  ;.  ..  .  , ;.,-•   ,  v;*:::::;:::!:::::;:^;::;:     -::::::::::::::::::::::::::';::"::!

 -   ,e-  * .  iv >i .. ^", *™1* ^ ,. *ซ,*-' J. K •--•'ซ>.-. ซ.l" -  !--!.•,**ป.-. ปซ•ซ•**••••• ป•••••• ••••••••••
"   • *.ซ    .,'•            u- ^ ป••'ป, u  i.  -..a*  •• ^-  p -" , -    .         .-.._,M,ปป.>.M."!..ป•ซ
s ,.-..:.-   .   •   ;.-....:,,,,. ^   .    . wt  ..  .....   ,      •  .     ,.,..ซ -i .._-... ~iซ* ...ป>
                                                                                                                                                    ... '. .... ., ป, .*••••••ซ••*••••••ป•
                                                                                                                                                   •*•••••ซ••••••••••••*•••*ป••••••••
                                                                                                                                                   •ซ•ซ•••ป*•ซ*••ป••**•ซ*••*•**ซปซ*••
                                                                                                                                                   • •••••••••^*v* *< *•- f • '** ••••••••••••

-------
     The ozone concentration is dependent on lamp  temperature,




lamp current, aperture setting, and flow rate.  The lamp temper-




ature is regulated by the thermal stability of the unit.  The




result of lamp current variation on the amount of  ozone generated




in individual calibration units was determined for a current range




of 5 to 18 mA, with a constant aperture setting of 0 (i.e., maximum




opening) and air flow rates of 150, 200 and 300 ml/min.  A typical




plot of this dependency is shown in Figure 15.  The dependency of




lamp current as a function of line voltage is shown in Fig. 16




for one flow rate and constant aperture setting.  At a flow rate




of 100 ml/min, ozone output deviation is approximately 1.5 percent




per volt change in AC line voltage.  These values are typical.




     Attenuating the photon flux incident on the quartz tube by




reducing the aperture diameter from 1.0 inch (micrometer set at 0),




to 0.5 inch  (micrometer set at 10.0) decreases the ozone output



of the calibration unit, as shown in Figure 14.




     The stability of the calibration unit is within the sensitivity




of the monitor and the experimental error of the calibration procedure.




Reproducibility error on the order of 6 percent was evident from the




calibration data, which also is within the determined repeatability




of the experimental procedure.




     Ozone concentration generated by the calibration unit at various




air flow rates,  source currents and aperture settings was determined




by the Neutral Buffered-Potassium Iodide Method and also by comparison




to a separate calibrated ozone source after the unit in test was




permanently installed in the monitor chassis.   The detailed calibration




technique is given in Appendix B.12.






-------
                                                             tO X 10 TO THE CENTIMETER   461313
                                                             KXZSCM                            Mt III U.I.*.
                                                                       KEUFFEL a CSSKK CO.
   ซ••••••*••••••.••ป
	••;••ปป*""ป*•ซ
S                                                                •**••••••••••••••• [*f ง•••••!• ป••••ป•• >-•••• ••••iซปiปpปปปซtrซ ••>ป•••• ป!••
                                                                • •••••••••••••••••••••••••••••MB •*ซซ••ซ•,•••ป•••*•{ป*ซ••*•••,••• Uปปซ*
                                                             • •••••••••ซ••••••*••*•ซ•••• •••••••••••••i ••**•••ซ•••••ป•*• MManlMM •*;
                                                             • •••*••*•*••••*•••••*••*•••*•ซ•***••••ป•*••**•ป•••*•••••*•*ซ••*•ป•••ซ••ป
                                                             ••••••••••!•••••••••ปป<••••••••••*••••••ซ••••••••••••••••••••••••••••••••
                                                                                                                                                                                                                     ••*••••-* *••••ปซซ•• ซ•ป•ป•••ซ•ป•
                                                                                                                                                                                                                     ซ••••*•••ซ•ซ•ซ••*ซ *ป*••ป*••••
                                                                                                                                                                                                                     !*••*ซ••?••*•ซ•ง••*>•!•••••••>*
 k**B •••••• ••ป••*ซ•••••••• •••••••*•••ซ&••
                                                                                                                                                        **••*••*••ปซ•*••••ป••ซ••**
                                                                                                                                                        **••ป••• •ซhซ*mซปUI •••••••
                                                                                                                                                        •*••ป******ซ***ป••••***ซซ
                                                                                                                                                        • *• *••*•* •ซซ•*••*••(••••••.•
                                                                                                                         ••r^**^2ff     w^"**"     'ฐ*            '
                                                                                                                         •*ซ••••*•••**••••*•••••*••••••••*•**•••*•
                                                                                                                         • •••••••••• m*ซ4          •••••••••*••***
                                                                                                                         ••>ป!•••••ป•••••ป••••!•••••fr-B*•*••*•*••
                                                                                                                                                                                                        ••*••ซ••••*ซ
                                                                                                                                                                                                        •*•••<* •••••ฃ
                                                                                                                                                                                                        *ป*ป••> WMMil
                                                               • ••>••>ซ••••••••••••••••••• •••ป••••••••!ป•ป•ป*••• gaaปซซMซปaปปซซaซp•ป•ป•ซ•**• ซ*ปซ•>ซ"ป 'AJปปB<ปซซซ••ปซปป•
                                                               • !ป••ป•• iMป*^iปปซg^ซซii**ป•ปซ•!•*K^piajaBaifcaaBJ^aaMMBassaa^a^s.^akiawiaaaiMaaaa.•*ซ'!ปซ -„ ,'   ^',ซ•••••••ซป•••*
                                                               •Miซaซfw*ซJUซUPBnaaaB4iiif*ซ*ซMซ>M*ปBAUBMmMซ*i|ซซBfซBc >•'%ซ*•-WBป>M|>ซซ* •ซ!ป••*•
                                                               •••ป*•ซ•••>• ซ•!•••*ซ•••••ป•••ซ*•••ซ•• kMCfa>ซniMBซซiiiBซ ป•••ซซป••••ซปป•••*•• •'. • ••  • i  *•**ป ,**••ซ •UBซll
                                                               ••*•••• •••*••ปป•••*•••• j           ป••*ป••••••ซ•••ซ*ซซซ••*ซ*•'• ปw, .. .ปซปtซ•ซ•ป•ป•*tl*llMIป•••••ป•
                                                               ••••••ซ•ซ••••••••••••*••••*ป••ซ•••*ซ••••••••••ซ••••*ป••ซป•••ซ••••ซ*•*••ป,•••ซ•••••ซ<ซ••ซ•••••*•••••••••B
                                                               • *••*•••••ปซซ••••••#*••••ซซ•ป•ป••ซ•.{ซ•••**••ป•••••••ป••••••ซ•ซ•••*•**ป .ซปป!.*ป•ปปปซปป•ซ•••ป•••ป••••ซ•••ซซ
                                                                                                     ••••*•**•••••••••
                                                                                                                   Wwwป-
                                                                                                                   ปซ•'
                                                                                                                                                         '••••••••••*~ *•>••ซ<
                                                                                                                                                         • i •*••••'•ซง•*•งป!
                                                                                                                                                         (••ปป•••• -••M|ซปปปซi
                                                                                                                                                         !•• •*'!••••••• ป•••ป•(
                                                                                                                                                                                                                                      • ^    tiBM^miiii ••••*•••
                                                                                                                                                                                                                                      ***ซซปซiซซ3nfvfrซaซซM
                                                                                  • •ซ•••ป•ซ It-',••ป••ซปปป-ป.
                                                                                  ซป•••••:• • -ซ*ซ•ซ••*•* ป• •ป•
                .* •* •*•
   n          Mil*** ป•*
•• ป•ปซ M   ซ•ปซ!*• ปrซ
ป•••    ••     M •• XV*
   *•          *• *• •*'
               ป" *• ซ•ซ
                -
           .
o  *iซ  a*,   *ซปซ          sซ   ป  "   4t                                              *    *ป ซซ     nซ               ***ป
3  an*   t*W   SBBป          ft*                                                      9    *tt *      ซ*               *ซปปซ
1ffl  *-*,  *0    ESปW        KM**-                                                      ซ    asHป55i      *E               ป**ซ
V  7^A       ?fr-^-^-.          Jซ                                                      *    •* *      X4T       *-**        ซ*
*  *ซซ   ซ*•ป•ป*          ป*           M                                            •• •       _!**        •••••*                 K
ii  s;tst       *eneป        ••**)ป                                                     __ *  ซ Mซ^Iซ                **st
*  Iซ*       *ซป*        V*  *•   **&*  #*                                             -'v, n  a      -a                ปisซr,j
ป  !•%•ซ•    ซ--J          n        h                                                •  •  •      ซ                ป•*•
ซ  ป•ป       *>!•ซ     &•&•*ป       Ml           •ซ•                                    (*.*•ป                )|ซIซ!JJ      *Hi     ••
^  E^ JL^       >4mซ          • w                        w                                  *  *   *• ป                •r^qa
*ซ• *   •ป  ปซป*          •ป  ปm*                                                      "  ซ      ฃ                'ปK*f
tf  ซ V       *••ป       ปซ *•                                                           ป  ซ      •                **p*
5  j ป       sซ*si          **                                                           ซ  "*     •                *ซป•
a  id 3>       )ซ*•          ••                                                           ป• ซ*  ปป ซ                ป••*
IF  W *       ซMH*          3*                                                           ปปซ*     •                ••ซซ
ซ  i a       ป&*a       -B*                                                           ซ••ป•ปป                *•••
9#ซ       li*ซiS          ซ                                                           ป••*•                ซ••ซ
fi  ป w       *.ป!•ป          5                                                           ป* ซ•     ซ                ปป••
ซ  • *       •••*          &4                        n               ซt               •••ปซ••      ••••
•*ซ       ป•*          *ซ•                                                      •ซ•*•                •ซ*•
I  • ป   M  •ป*ป          •                                                           ปป ••     p                *•••
*  ฃ ซ       !•••          •                                                           ปซป8ซซ                ••ป>
•  • *       *4*n  ••       •                                                           *• ซป     •ซ               ซ••ซ
*  ซ s       *s^>ซ          •   *                                                      s* ซซ     ••               ปซ•ซ
aS*.SijHi5ซ!'*          H*,*J>           tyf       •                               ** BlTrf    ••          ซ•••••
ซ  (i ซ   SB ••*••         * *   *s   *^      *    SS   R    ซi                               *ป *ป    ซ"ซ "       ซ* •ป•ป•••
•ปป   •  •••ซซ•      *ซ•*   ••   ••-•    ••    ซ•  ซ*    W>*          •*       *          ป*•••ป  ซ*• ซ       *•ป••*!ป*ซ
>  • •  •• •••*••  E     •ป*•*   •*••••     i*    ••••    aซa          *e       *         **"*!"  aim •       ซ••ซซ
                                   **•••*•• *nซ**B*ซaiปkli .**••• •*•••••ซ••••••••••••••••*••* •*ซ••••••• <ซซ•••••••••••••ซ••• •**ปซ***••ปป*ป:ปป J a (Bfjat*ซปซปซ** ซ•••ป ปซKซ**a*.. ii
                                   • ป•<•• •*•*•••ซ•• •••*)! ••••ซ •ซ•ซ•ซป••• ••••Vailll* ••*ซ*)>••ซ••••*'•ซ••*-** •ซ**••***" •••01
ป••*••ป•ป•••ซปป ••ซป•
ป•*•**••••ง•ซ••&*•*•
          n              M
          •              E
          ;    „        ;
        i :  :Ji   ; ii :::
S ii ซH :  insrซ-K :si ^-^
  ;   i          j     t   *  """";       s                    i       -*          i       i
  :                  •*   - ::H                           -s       5                 :

  s             "        S3 •*••                  ****ซ•                 m
  w                         ^B it if it                             m       *                 *
  ;                    ป  J;:;!.                                    ซ                 ;
f :          i Siii;  ฃ   HUH*  ..:                       Hi., i :        u       I:   III
h-ij      uJsBif"  jntsa    ri 5    !    -  S     U:!!H!I !     :   HH     s:   ":
                                                                                                                                           •*
                                                                                                                                           ป•
                                                                                                                                           ;          •
                                                                                                                                       ,    L.       ::: 1
                                                                                                                                           :      HIM ^

                                                                                                                                        J
                                                                                                                                5    „ :  :
                                                                                                                        i5 1.  r    : s
                                                                                                                                                                                                                 ifttfPซ4Bvp*ฃ#4IIV •ป VtfVBCd •**

-------
                             ••{•••••••••••••••••••••••^•••^•••••••••••••••••••••••f ••ซ•••ซ••••ป••••*ซ••*••ปซ•
                             |gfj|ซ••ซ•••• fifiปiiปซ*ซปIซซซft          ซป•••••• ••••••••••(•••••••••••••••••••Iป•ซ
                             fillliMi pปiiiปj*M ••*>••••*•ซ iftBP'MCBB^
                             |IlijiMM*iซซM
                             •••••••• •••••MซM|pNipp>MPpp ••••••ป PP *••••*••ซ• *••••ป•••ป{••*• •ป*•! •••••• ซปซIBPซ
                                                                                           [•ป**ปป*ปtป*ป
                    iiniHiliiiiEiซipa•••ซซ•••••ป<•••*nซซ                      ป•••••ง•• ••••••••!
                    iiBlซiijiSiif•!•ง•• ••ป•••**•ฃ*ป•••••"•"•*•"*••*•*"•••••••ซ•••••••••••••••••••••••ซ

                                   :
BttHffiffBwn         mffl} SffiffiBfflm
                                                               B • ••••PPPBPHPBMPH Pr *ป••ป.


                                                               ••::••••:::::!:::'!::::!:

                                                               PPPPVJMP* IPPB BPPB* ••••(•••••
                                                               ••••i>ปปซ*Bป*PF*•ง••ซ••ป*
                                                               •••••••••••Pซrj*||pซBlaap
                                                               • •••PPPPBPBPV Jป*MPปPซBซB
                                                               PPBPVPPซซPP*dPttPBPAPPPPPP
                                                           • •••• •••••••••I •••••••••S                *•! ••*ป(ซ PซPซป ป•••••••• BliPP ซ•••">ป •.•••ป**!ปซ*
                                                           apVPซ*ซpBp*BBBi~pappi|pBA          n**vซ ซ•••ซ••ซซป<*•••**•*•• ***Bซ ••••*•>•••**•••*ซ'ซป••ซ•ป•••*ปป*•••*••*•**ซ
                                      Miiiiiitli
                                                          S*"*1**1'*"" J ••••ป• P •• Bp
                                                                             •••• P •••ป I* •••• ppปBiปtปP*P
                                                                                                                       • ••••*ซp*PซPPPBPaPPPBPPPPHPVPVBVkVPBVB Pป

-------
     A calibrated ozone generator identical to the installed unit,

preceded by an ozone-moisture trap and a mass flow meter, was attached

to the sampling port of the ozc*ne monitor being tested (see Fig. 17).

Operating the monitor in the sample mode provided comparison data,

which was indicative of the final calibration for the installed unit

at various aperture settings.  The stability and reproducibility

of the individual unit was also observed.


     3.5  Timing and Control Unit

          The timing and control unit provides selectable timing cycles

for the sample air and calibrate solenoids and the calibrate lamp

shutter.  A circuit diagram for the timing and control unit is shown

in Figure 18.  There are two automatic modes and three manual modes of

operation which are push button selectable from the front panel of the

instrument.  The manual modes are as follows:

           CONT MEAS - continuous operation in the measure mode; i.e.,
                       the sample air solenoid open and the calibration
                       lamp shutter closed.

           CONT CAL  - continuous operation in the calibrate mode; i.e.,
                       the calibrate air solenoid open and the calibrate
                       lamp shutter open.

           CONT PURGE - continuous operation with the calibrate solenoid
                        open, but with the calibrate shutter closed.

           EXT        - remote switch for external control of mode.

The two automatic modes of operation are essentially as follows:

           SAMP I  - 5 min cycle consisting of 75 sec measure, 75 sec
                     purge, 75 sec calibrate, and 75 sec purge.

           SAMP II - operation in the SAMP I mode continuously except
                     for the 30 min PURGE mode every 12 hours.

-------
      CONTROL

117 V 	ฉ—+
                                            -*•	v._

                                       	1  / ) MA METER
                                           XC~X
AMB. AIR
 INLET
       SCRUBBER
           MASS FLOW
             METER
CALIBRATED
  OZONE
  SOURCE
                                                        MEAS.
                                                        PORT
OZONE
METER
                                                     12 v.d.c.
             • EXHAUST
              PORT
                                                                  ANALOG
                                                                 RECORDER
                  FIGURE 17.   CALIBRATION UNIT,  IN SITU CALIBRATION

-------
_KOOE__COK1TOOL
       I
                                 L_	
                         10
                                              K3?.
                                                                                               PURGE
                                                                                                      JฉHTS
rr —

T^i— — i
!
f< 	 .
1



^~^
i
PURGE
TIMER

	 K!
^

	 /"



























































i
!
i
!


















j..

8
^ ^
_L
"








                                                                                                   MEAS.
                                                                                              "••  SOL. VALVE
                                                                                                   CAL.
                                                                                              -'  SOL. VALVE
                                                                                                   CAL.
                                                                                                 SHUTTER
                                                                                                 MEAS.

                                                                                                 CAL.
        ^•S/H
}
                                                                                                  EVENT
                                                                                                  MARKER
            n a 12
NOTE: SYSTEM  SHOWW IW
      "SAMP. 1 - MEASURE  MODE" CONFIGURATION .
XFMR
1 	
" r
-o-
^
O- .
Pi

~<'rV/1 1 120 VAC /60 HZ
17-,
..,o SHUTTER
IBlO INTERLOCK
I@Q CIRCUIT

                                                                            BUZZER

-------
The indicated mode times for the automatic modes are those initially

set into each unit; these are adjustable and may be set for longer

or shorter periods of time.  The timer adjustment procedure is given

in Appendix B.6.

     A 12 VDC power supply, located on the main chassis supplies

the voltages for the solenoids, calibrate lamp shutter relays and

indicator lamps.

     Event marker connections on the rear panel are connected to

relay contacts which are closed coincident with the actuation of

the calibrate air solenoid and the sample periods for the measure,

calibrate and purge S/H circuits.

     The AC voltage supply and buzzer for the PM tube shutter inter-

lock circuit are located on the timer chassis.  Details of the inter-

lock circuit are given in Section 3.3.


     3.6  Linear Amplifier

          3.6.1  Description — The amplifier performs the function of

raising the photomultiplier tube anode current from the lowest expected

                          -9
values, on the order of 10   amps, to the desired 0-1 volt output.  A

circuit diagram of the amplifier subsystem is given in Figure 19.  The

amplifier is basically a calibrated current meter incorporating a test

position and five linear ranges.  In addition, certain other desirable

functions are provided such as an zero offset, span and time constant

control.   A summary of the amplifier characteristics is given below:

          Sensitivity:  Linear - 10~9 amps, FS     >
                                 10~8 amps, FS
                                 10~7 amps, FS
                                 10~6 amps, FS
                                 10~5 amps, FS

-------
                                                       TOWER SUPPLY
LIWE&O  OKHJT
 AMPLIFIER
19
                                           MOTE= ALL RESISTORS 1% B1ETAL RLM, EKCEPT
                                               K-l%  CAR30M
                                               ปK-GLASS EWCLOSEO

-------
          Gain Accuracy - 1%



          Input Impedance - 10  ohm

                            (effective load impedance for PM tube)



          Noise           - 10 pV p-p, .01 to 1 Hz (time constant-normal)



          Output          - 0-1 volt, recorder output jack

                            (5 mA max current)



          Drift           - Less than 1% after 1 hr warm-up



          Time Constant   - Normal (see Fig. 21)

                            1 sec, 2 sec, 4 sec

                                   _Q

          Zero Offset     - 0 to 10   amps



     The input current (PM tube anode current) is passed through a



10   ohm voltage divider.  This voltage divider not only provides a



known load resistance but on the two least sensitive ranges it



provides voltage division of 10 or 100 to prevent overload of the



input stages of the amplifier.



     The voltage drop across the divider is then sensed by a



varactor bridge operational amplifier.  This particular unit, the


                                                              -14
Analog Devices 311J, was selected for its low bias current (10    amp),



low bias current drift (10    amp/ฐC) and high input impedance


   14
(10   ohm).  This amplifier boosts the lower level signals to a



usable level as well  as  isolating divider from  any loading



effects.



     The calibration curves for the linear scales are shown in Figure 20.



The signal is fed into the output amplifier through a ten-turn linear



potentiometer, which provides a continuously variable span control.



Capacitors are placed across the feedback resistor in the output



amplifier to provide filtering to reduce high frequency noise and



smooth the output.  A four-position switch selects the proper capacitor

-------
      mmm+mmmmmmmmmmmmmmmmmmmft&                         *>•••••••••••••••••• •••••••••••••••••••••••••••••••^••wtf •*••••••*••*•*•••••••••ป•••*••••••*••ป•••••*
      ••••••••••••• •••••ป•••••ป*••••ซ•ป•••••••••••••ป••••• ••*••ป>•••ป•••ป••••!••ซ•••ป••••••••••• *••••••••••ป•••••••••••••••••••••••••••ป•••ป•••ป••••••
      *ซ•••••••••ป••••••*•••*•••ป••ซ••••••••••••••••••••ป••••••••••ป••••••*•••mmmmmmmmmt•••••••••••••PB••••••••••ซ•••••••••••••••••••ป••••••••••*•••••••••
      ••**•••••••ง•••••V^                  '*••••••••ป•••*••;ป•ป••••••••••••••••••••*•••ซ•••ซ••••••*•••ป•••ซ••••
      ซ•ป•••••••**•••ป•••••••••••ป••••ซปซ*••••••••••••••••ซ•••••••ป•••ป•ซ*"••ซ••••••••ป••••ป•••••ป••••••ป•••             Ml*ซ•••!
                          !•••••••ซ•••ซ••••••••••••••••••••••••ป•••••••••(

• •••••••••••••••••• ••ซ•!)           ปป••••••••••••*••••••••••••••ซ••••• ••ซ!
        ซ••••••••••PIซIB•••*•••••>•

-------
for a time constant of normal, 1, 2, or 4 seconds.  The amplifier



band-width for the normal position of the time constant control



is shown in Figure 21.



     The effectiveness of the filters were checked by applying



step functions to the input and recording the output either on



a scope with camera or on a strip chart recorder.  These are



shown in the scope photographs in Fig. 22 for time constant set-



tings of normal and 1 sec, respectively.  The upper trace is the



output response, the input waveform is shown in the lower trace.



The horizontal sweep and vertical gain are identical in each



illustration.



     The output is brought out of the amplifier box through two



BNC connectors connected in parallel.  The front panel meter is



connected to one and a rear panel recorder output jack is connected



to the other.



     Zero offset or base line adjustment is provided by a bucking



voltage fed into the top of the 10  ohm PM tube anode load.  This



is controlled by a 10 turn linear potentiometer and is adjusted



for zero offset at zero reading on the dial, and provides linear


               —8
offset up to 10   amps at a dial setting of 100.



     The unit is completely self contained.  An Analog Devices



100 mA + 15 volt power supply allows the unit to be operated



directly from 115 VAC.  It is enclosed in an aluminum box with



removable side plates and switch shafts long enough to reach through



the front panel.

-------
• ^••••• ••••JUMIMMIMMMMป
  • •mi 4ttiil>ll< ซ*g|llil i
  •iiiiiiiilMiliiii JilMC '
••iiiiiiuiiitiiiMi iiiiiiiili
• •••••**I!Wป*ซIIซB*ซ ••••tซpซSJซปซ*HIซ#*ปซlซปซปl.ซlซป*lซซWซปซปซซ*ปปซป VซB*Xซ'ปซปซBซป>*>^
• •Mllili:tซMII>lll ••••ปIIUBIMIIplliaซIlilllllซll>illllปBMH(lHIHIIIIItaBMIHIII
                                                     MMVtu.. >••ป•ซ* I. :-BK*ป'ซVปJMซปMirซMซปtBI
                            1|M^  REPRODUCIBLE
               iliiliiii,''".,,wlttsMffliii
Figure 21.   Frequency Response of Amplifier for Various Gain Settings
ri 11 ii inn i 'in nil 111 :!"TTTr.'pi i f "nrnrrpTrrnr i rTirayTrn':iirn'!Tir'!TTi :~rrmn THI I'I.TTHI nil ll MlMi.

-------
         IV




OUTPUT    0



        -IV








        -IV




  INPUT    0



         IV
i.i    1.1   1.1
                  10      20     30      40




                             (a) T = NORMAL
                   50
                       SEC.
         IV




OUTPUT    0




        -IV
        -IV
 INPUT
         IV
                                          _J	1	I
                  10      20     30      40     50




                             (b)  T = 1 SECOND
                                                     SEC.
                       ,-7
Figure 22-  Amplifier 10   Amp Full Scale,  Output Sensitivity XI

-------
     Standard maintenance and calibration procedures for the amplifier




are given in Appendix B.7.






     3.7  Signal Conditioner




          The purpose of the signal conditioning unit is to sample




the amplifier output signal from the amplifier at the appropriate




points in time; i.e., in the measure, calibrate and purge modes,




and properly combine these signals in such a way that the resulting




signal is a smoothed continuous value which has been corrected for




base-line and calibration values.




     A diagram showing the signal and sampling time-relationship




is presented in Fig. 23, a functional circuit diagram is shown




in Figure 24.  The normal time-sampled signal, consisting of 75 sec




calibrate, 75 sec purge, 75 sec measure and 75 sec purge constitute




one complete sampling cycle.  Timing signals for the sample and




hold amplifiers are shown as the control inputs - calibrate, purge




and measure, respectively.   These control signals initiate the




reset condition in which the s&mple and hold amplifier acquires the




new input signal and then holds this signal at its output until the




next reset signal.




     As illustrated in the functional diagram the purge signal is




subtracted from the calibrate and measure signals and the resultant




signals are presented to the inputs of the analog divider as c-p




and m-p, respectively.  The subtraction is accomplished in the




differential amplifiers.  The output of the analog divider is,

-------
(a)
               1.0V
Ozone
Monitor
Normal
Output
Signal
                              cal

                               *
                                                           mea
(b)
(c)
             t
                     Calib.
 Control     I
 Mode Signals^ Purge
             t
                     Meas.
 Output of
 Signal
   S/H
 Amplifier
                     Calib.
                     Meas.
                     Purge
                              -ฃv
                                                                                	n.

-------
(SI







INPUT
SIGNAL























SAMPLE & HOLD
(CALIB)
MODE _}
CONTROL



'JAMPT E & HOLD

(PURGE)
MODE \
CONTROL

SAMPLE -& HOLD
(MEASURE)
MODE 	 J
CONTROL
-m






-P





-c













^






"^x.
DIFFN>m~P
s^
+ AMEX
^ * "
DIVIDER 	 ป-


+ DIFFX
s' c~p
/^


CONSTRAINTS : y>0
lซl<
                                                                                                         -10
                                                                                                             c-p

-------
 (as one would expect it to do due to disk delay) then, for a given




 concentration of ozone, the measure signal is automatically corrected.




 That is, initially a calibration is made and a given concentration




 of ozone produces a certain output signal; as conditions change,




 such as the decay of the disk then the output signal is corrected




 to produce the same value for a given concentration of ozone.




     An output amplifier and filter combination is used to smooth




 the discontinuities in the signals due to the discreet time




 sampling characteristics of the sample and hold amplifier.  The




 output is a continuous signal representing the ozone concentration




 for a fraction of each 5 min cycle.




     The detailed circuit diagram for the signal conditioner unit




 is shown in Figure 25.  The normal output signal from the ozone




 meter; i.e., the cyclic signal shown in Fig. 23(a), is amplified




 by the X10 input amplifier and this signal is presented simultan-




 eously to all three inputs of the sample and hold (S/H) amplifiers.




The S/H mode control signals are obtained from the timing & control




unit.   The output of the S/H amplifier is -10 V for a normal input




 to the system of 1.0 V.  These S/H signals are presented to the




respective inputs of the two difference amplifiers.  Since one of




 the constraints of the analog divider network is that |m-p|_<|c-p|




a *10 network was inserted in the output of the m-p difference




amplifier.   This accomplishes several things; it satisfies the




constraint up to an effective value of mป10(c) which allows one to




keep the calibration signal at a reasonable level and it cancels

-------
               INPUT
             AMPLIFIER
               XIO

               t
       10K
     -AY
&*L
              t   -H5V
               • i'.T   I
            •AD183J
INPUT
     ฐ^jlYc
       ^ฑ-    4- -15V
          ^IOK
          1
  DC POWER SUPPLY
              RD
                 +157
                 . -15V
                              SAMPLE & EOLD CIRCUITS
                                     K(-t)
                                      L  nf

2
! "
1 4
BB4013/25
-t- C - 3
1 I J 1

—


^" MEAS
              +15V - -15V
              +15V '  -15V
                       I ULf
                      —-It—
                           I—J2 BB4013/25
                             L
                 + 9-3
                TTi
                             +15V  " -15V +15V
                                   f^lU-L-.
                                                                   DIFFERENCE AMPLIFIERS
                                                                       IOK

2 BB4<
+ <
Srf

1 4
513/25
5
: - 3



f




                                                    -O CA1
                                     AMPHENOL
rfriT
i
L
GR
VI
GR
OR
BK

TIMER UNIT
MODE CONTROL SIGNALS
ZERO VOLTS - SAMPLE MODE
+15 VOLTS - HOLD MODE
 OUTPUT
AMPLIFIER
   10 Uf
    II-

-------
the factor of 10 existing in the analog divider output.   The output




amplifier converts to a positive going signal and provides for some




smoothing of the inherent step function response of the S/H amplifiers




during the updating period.

-------
                               APPENDICES
Appendix

   A.     INSTALLATION AND OPERATION

          A.I  Preliminary Set-Up Procedure
          A.2  Turn-on Procedure
          A. 3  Operational Checks and Adjustments

   B.     MAINTENANCE, CALIBRATION AND ADJUSTMENT PROCEDURES     B-l

          B.I  Main Chassis Removal                              B-l
          B.2  Disc Replacement                                  B-l
          B.3  Flow Rate Adjustment and Check                    B-2
          B.4  Filter Maintenance                                B-4
          B.5  UV Lamp Replacement                               B-4
          B.6  Timer Adjustment                                  B-6
          B.7  Amplifier Calibration                             B-8
          B.8  Signal Conditioner Calibration Check              B-9
          B.9  Calibration Range Change                          B-9
         B.10  PM Tube Replacement                               B-ll
         B.ll  Calibration Procedures for the Calibration Unit   B-12
         B.12  Chemiluminescent Disk Preparation                 B-17

   C.     PARTS LIST                                             C-l

          C.I  Detector Assembly                                 C-2
          C.2  Calibration Assembly                              C-2
          C.3  Plumbing Assembly                                 C-3
          C.4  Timing Control                                    C-3
          C.5  Amplifier Assembly                                C-4
          C.6  Signal Conditioning Unit                          C-5
          C. 7  Chassis and Cabinet                               C-5
          C.8  Interlock Circuit                                 C-6
          C.9  AC Power                                          C-6

   D.     SUBSYSTEM MANUALS                                      D-l


-------
                               APPENDIX A

                       INSTALLATION AND OPERATION

                                  r
     A.I  Preliminary Set-up Procedure

(1)  Access to Main Chassis

     (a)  Remove screws from front panel.

     (b)  Check for freedom of movement of air inlet lines signal cable,
          and power cable at rear of unit, or disconnect as appropriate.

     (c)  Slide chassis out until it stops.

               NOTE:  A safety chain restrains the chassis from
                      extending beyond the first stop.   This should
                      not be removed unless  chassis is  to be removed
                      completely from cabinet.

(2)  Install Photomultiplier Tube

     (a)  Install the PM tube in the PM tube housing and mount on
          detector unit.

               NOTE:  Care should be taken not to expose the PM
                      tube to strong light;  excessive exposure
                      requires approximately 24-48 hours for dark
                      current to decay to  normal value.

(3)  Attach the HV and signal coax cables  to the PM tube housing.
                                                    *
(4)  Remove shunt wire from output meter terminals.  (This shunt wire
     is a precautionary measure taken in shipping to prevent damage).

(5)  Visually Inspect unit for any obvious damage.

(6)  Slide main chassis back in place.

(7)  Install chemiluminescent disc (Appendix B.2).


            CAUTION:  The instrument is designed for normal
                      operation with both  front and rear panels in
                      the closed condition.   Opening either end will
                      result in a temperature rise in certain components,
                      Complete removal of  chassis is satisfactory pro-
                      viding ventilation below chassis  is provided.

-------
     A. 2  Turn-On Procedure

(1)  Close shutter - full closure indicated by AMBER Lamp.

(2)  Set SENSITIVITY to 5.

(3)  Set SPAN to 0.

(4)  Set TIME CONSTANT to 1 sec.

(5)  Adjust HV to recommended value for gain of 200 A/L (see specification
     sheet).

(6)  REPLACE HV Power Supply Cover.

(7)  Turn AC POWER switch to ON.

               NOTE:  Initial operation procedures should include
                      purging of the air intake lines with high
                      levels of ozone to destroy contaminants.


     A.3  Operational Adjustments/Checks

(1)  Adjust LAMP CURRENT to 17.5 mA.

(2)  Set FLOW to 200 ml/min.   (see spec sheet for proper, flowmeter
                                settings)  (See Section B.3 for
                                procedure)
(3)  Check dark current - note value.

(4)  Open shutter - full open indicated by  RED Lamp.

(5)  Switch to GAL mode - note value.

(6)  Switch to desired operating mode.-

(7)  Set amplifier controls to desired range and time constant.

-------
                               APPENDIX B

           MAINTENANCE, CALIBRATION AND ADJUSTMENT PROCEDURES


     B.1  Main Chassis Removal

(1)  Remove screws from front panel.

(2)  Slide chassis out a few inches and disconnect the safety chain.

(3)  Disconnect HV power supply AC cord and output cable, and fan power
     cord.

(4)  Slide chassis out until safety catches on chassis slides stop move-
     ment.  (The weight of the instrument should be supported by the
     operator.)

(5)  Depress safety catches on slides and remove chassis from cabinet.

          NOTE:  When operating unit out of cabinet, raise chassis
                 l"-2" off bench for increased component ventilation.


     B.2  Disc Replacement

(1)  Move shutter control to CLOSED position.   CLOSED indicator light
     mus t be on.

(?.)  Open disc access door and lower disc holder by turning knob counter-
     clockwise until holder stops.   Lift holder enough to clear locating
     pins and pull unit forward.  Be careful not to drag against top
     plate of unit.

(3)  Replace old disc with new o;.ie making note of elapsed time.

(4)  Place holder back in unit using two locating pins on bottom plate
     to correctly position holder.   Be sure Neoprene outlet hose is not
     pinched behind holder, and 0-ring is properly seated.

(5)  Turn knob clockwise until holder is firmly seated.   If possible, the
     system should be checked for leaks each time the disc is changed to
     avoid erroneous data.

-------
     B.3  Flow Rate Adjustment and Check

(1)   The proper flow rate is 200 ml/min.

(2)   Adjustment of the air sampling rate is accomplished with the fine
     control valve (front panel mounted) and the coarse valve (located
     just behind the front panel).

(3)   Flow rates at the ambient air sampling port should be checked
     periodically, either with a mass flow meter device or with a
     rotameter capable of measuring 200 ml/min with an accuracy of + 2%.
     It  is recommended that where an external flow meter is available
     that it be used to make final adjustments to the air flow.

(4)   A calibration curve for the flowmeter is shown in Fig. B-l.

-------
 ••{••MB*i••••••If•ป'•••••••ป•••*
 *•••••*••••••••••••••*••ป•••ป•••
                               Hj
                               •i
                               ••

                                                                              K. **rj, 1ซ ซ s *.ป•• i: >ซซp J -! • •  • '. :.    "i .- i.  -•..->-.• p*.?. - ป• •- ."-.,ป, ,• -. ,,J,., - , ซ^.
                                                                              rป**ซปปป**1ปliปป'*ซป1ซ-**ซnป**ซmซซป"*ปปซ"***Pปnปป"ปซซปปป*ปซซซ*ปซ
                                                                              ::!::::::;•:::•:•:•:::::::::.—:::::::::::?::::::::::•;::::::
                                                                              :-:::     ;::::::::::::::::::::::::::::::::::::::::::::::::
                                                                              K - r. j*J.-. I *ซ*i.^*=. *ป •]=->.,,!  ..:. -  .   .:-->..•',.•;-.•ป.,* -*,iซซปซs'Kii. ,*ป*.* nwa
                                                                              :::::::::ป::::::::!:::::::::::::r;^irrv:::r:5:?::::::-s;::;
                                                                              • •••••••••••••ซ••• •••••••••*••ซ•ป---  * * . ป.* .-.ป...., n *„..,. .-I,,,
                                                                              i!IHi:5llซHIIซlซH!iปHH!!5=:i:illll!ll!liill!lllil[l
                                                                              :::::::::::::::::::::::::::::::     ,:;.;;;: .:;;::::ซ:::::
i?::-::::::::::;:;::::::::::
• •••.•SIB ••••••*•••*•*•••ซ••• ซ
••••-.*•*••••ป•••••••••••••••
                            ?
• ••••••••••••••••I *•••••*•!••*
'•••••••••••••••••ซ•••••••••••
                                                                                    I:::::::;::::::::::::::
                                                                                    ปป.>K.Iซ.ป.>ซII.<..>>.B^~.^
                                                                                     , -:* u, ^ ซ . J, ,  .  ..   >  „
                                                                                    ••••Iป.ป-ซ3•••ซ••••••••
                                                                                    ป t™,ui,. ,^ .- , ,,-, ,,r ป. , ..ซ
                                                                                    *ป* lean •*•ป••••••{!••ป••

-------
     B.4   Filter Maintenance

 (1)  Filter A  (see  Fig. B-2) is a millipore aerosol filter which has been
     modified  and includes  filter paper, glass wool, activated charcoal,
     and manganese  dioxide.  The filter is prepared by placing a disk of
     Whatman #541 filter paper in the base of the aerosol filter.  A
     layer of  activated charcoal is placed up to the first lip and is
     covered by a second disk of filter paper.  The remainder of the
     aerosol filter is filled with glass wool.  Manganese dioxide is
     then  sprinkled and allowed to sift into the glass wool.  The aerosol
     filter is reassembled  by fitting the second section back into the
     first section.  To insure that the filter is air tight, cover with
     tape  or seal the joint using acetone.  Under normal monitoring con-
     ditions this filter should quantitatively destroy ozone for a period
     of at least 12 months; however, replacement at shorter intervals is
     recommended when high  concentrations of ozone are measured routinely.

 (2)  Filter B  is a  disposable millipore aerosol filter filled with glass
     wool.  The glass wool  should be replaced as needed.  Visual inspec-
     tion  at 6 month intervals should suffice.
     B.5  UV Lamp Replacement

(1)  Turn AC power OFF.

(2)  Disconnect lamp from transformer with in-line connector.  (CAUTION:
     Power must be OFF before disconnecting lamp from transformer.)

(3)  Remove top plate.

(4)  Loosen Allen set-screw holding UV lamp.

(5)  Note position of tip on UV lamp with respect to end wall.  Slide
     lamp out towards rear of unit, removing shield at same time.

(6)  Install new lamp with tip at same position as old lamp, installing
     the shield as the lamp is slid into position.

(7)  Tighten Allen .set-screw just enough to hold lamp in position, as
     excessive tightening could damage the lamp insulator.

(8)  Opening in the  shield should be in position for maximum light through
     aperture.

(9)  Install top plate making sure micrometer shaft is in aperature arm
     socket  and connect in-line power connector.

(10)  Turn on AC power.

(11)  Calibrate  output against known ozone concentration,  or use standard
     KI technique.

-------
AIR FLOW
                                     FILTER PAPER
                           ACTIVATED
                           CHARCOAL
           GLASS WOOL
          POWDERED WITH
    FIGURE B-2.  FILTER "A" - CUTAWAY DRAWING

-------
B.6  Timer Adjustment

     B. 6.1  Main Timer and Cycle Control

       (1)  The basic stopping time for the programming switch is
            controlled by the small continuous running switch.
            The 1 rpm allows for selection of basic stopping times
            from 1 sec intervals' to 60 sec intervals.

            The unit is presently adjusted for 5 sec intervals.

       (2)  The 7 channel programming switch is stopped at 5 sec
            intervals as indicated in (1).  Each function is color
            coded as follows:

              Red activators - purge mode (25 sec—twice per cycle)
              White activators - calibrate mode (25 sec—once per
                          cycle)

              No activators - measure mode (25 sec—once per cycle)
              Single blue activator - sample-mode (5 sec) for
                          respective S/H amplifiers.
              Additional channels used for mode control lights and
                          event marker.

       (3)  The activation time for the respective modes may be
            readily changed by means of adding or deleting the
            activators in the respective channels.

-------
B.6.2  Purge Time

  (1)  The purge timing  cam is  mounted on a shaft by means of  a
       heavy duty  friction  unit which allows for each finger
       adjustment  of  the timing sequence,,
(2)   The cam may be set  for  "on-time"  of from 2% to
     total overall time  cycle.
                                                           of the
  (3)  The cam opening  or  "on-time" may be adjusted by loosening
       the cam screw  and turning the movable cam to the required
       degree of opening and then re-tightening the screw.

  (4)  See Fig. B-3 for identification of components.
                          CAM
                        ADJUSTMENT.
                          SCREW
                     8VNC. MOTOR
                                     OPEN
                                    {COMMON

                       LINE

                   WIRING DIAGRAM ONE OB MORE CAMS
                 FIGURE B-3.   TIMER ADJUSTMENT

-------
     B.7  Amplifier Calibration

          It is not necessary to remove the amplifier from the unit

for this calibration.  Simply disconnect the input and attach a volt-

meter to the output.

          Equipment needed - 10 mV and 10 V source
                           - Voltmeter - HP-425, or equivalent
                           - Oscilloscope - Tektronix 547, or equivalent.

          All adjustments are located on top of the unit and labeled.

(1)  Turn the unit on and allow 30 min for amplifier to stabilize.   _
     Leave the input disconnected.  Then, turn the sensitivity to 10  ,
     F.S. output to XI and time constant to NORM.  Adjust Rl for zero
     output.
                                                        —5       _Q
(2)  Turn the sensitivity control alternately between 10   and 10
     while adjusting Rl for a minimum change in output between switch
     settings.

(3)  Set the sensitivity control at 10   and adjust R2 for zero output.
          Amplifier Disassembly - To remove the amplifier from the
          unit, remove the three knobs and three screws from the
          front panel.  Next disconnect the three BNC connectors
          from the top of the amplifier and unplug the power cord.
          The back of the amplifier can now be raised and the unit
          pulled out over the timing chassis.

-------
     B. 8  Signal Conditioner Calibration Check

(1)  Disconnect input to signal conditioner and reconnect a 1.0 V dc
     signal.

(2)  Disable timer by removing AC power.  Rotate timer to desired
     position manually.

(3)  Measure with DVM the voltage at each sample and hold output;
     should be 10.0 V + 1%.

(4)  Check the * 10 at the output of the upper difference amplifier.

(5)  Under normal operating conditions, read the respective S/H
     signals and calculate the (m-p)/(c-p)  ratio.   Measure the output
     signal and compare to calculated value.
     B.9  Calibration Range Change

(1)   Turn AC power off.

(2)   Disconnect safety chain from rear of chassis and pull chassis
     forward until calibration unit is completely exposed.

(3)   Remove top plate.

(4)   See Fig.  B-4 for following steps:

     a.   To adjust for lower calibration range,  set micrometer at 10,
         loosen the two  4-40 retaining screws and rotate aperture cover
         counter-clockwise until minimum aperture opening is obtained.

     b.   To adjust for higher calibration range, set micrometer at 0,
         loosen the two  4-40 retaining screws and rotate aperture cover
         clockwise until maximum aperture opening is obtained.

(5)   Tighten the aperture cover retaining screws and install top plate,
     making sure micrometer shaft is in aperture arm socket.

(6)   Calibrate output against known ozone concentration, or use standard
     KI  technique.

-------
     ALUMINUM
     REFLECTOR

         \
PEN RAY
UV LAMP
VARIABLE DIA.
 APERTURE
                                  MICROMETER
                       tunas
                        Retaining
                         Screws

                                            QUARTZ TUBE
                  Aperture Adjustment Range

                  Small:  .047" to 0.5" dia.
                  Large:  .5" to .984" dia.
            FIGURE B-4.  CALIBRATION UNIT (side view)

-------
     B.10 Replacement of PM Tube

(1)   Turn off AC power.

(2)   Remove screws from front panel and slide the chassis out until it
     stopso

(3)   Disconnect the HV power supply and signal cables from the top of the
     PM tube housing.

         ! CAUTION - When handling PM tubes,  avoid exposing the tube    '
         '           to light.  This will insure a normal dark current  :
         |           level for the tube.  If  exposed,  allow 24-48 hours :
         j           for return to normal level.                        :

(4)   Remove three 8-32 screws from the PM housing flange and lift the
     complete housing assembly off the detector unit,,

(5)   Remove three 6-32 screws from the top of the housing (evenly spread
     on outside rim) and remove the outer housing.

(6)   Remove the Mu-metal shield from the PM tube and  then unplug the
     PM tube.

(7)   Install the new tube9  reversing the above procedure, being careful
     not to dislodge the metal grounding braid near the end of the
     Mu-metal shield nearest the cathode at  the PM tube.

          NOTE;  Light leaks will show up as excessive dark current.
                 Care should be exercised in assemblying the housing.
                 (a)  It may be necessary in some case to use black
                 tape to seal the joints of  the housing around the
                 BNC plug end,
                 (b)  An 0-ring in the mounting flange of the tube
                 provides an adequate light-tight seal for the open end
                 of the housing.
                 (c)  A check should be made around the PM tube shutter
                 to determine the susceptibility to light leakage,

-------
    B-ll  Calibration Procedures for the Calibration Unit
          Apparatus - The unit  in  test was placed  in  line with  an ozone-




moisture  trap of Drierite and No.  40 mesh charcoal, a mass  flowmeter,




two  25 ml graduated all glass inpingers  (open end  nozzle type) ,  an air




flow control valve and air pump  (refer to Fig.B-5).   Quartz tube to




impinger  connection was accomplished by means of a teflon reducing




union, glass tubing and aground  glass joint, slightly lubricated with




Kel-F No. 90 stopcock grease.  A 12 VDC power supply  was used to




activate  the internal shutter and  the UV lamp current was adjusted by




means of  an autotransformer and  a  Pen Ray 10 mA transformer.  All




tests were timed with an electronic timer and AC power was  regulated




at 117 volts.




          Air flow was metered with a Hastings mass flowmeter (a ther-




moelectric type sensor) with a range of 0 to 300 standard ml/min,




readable  to 1.5 ml/min and stable within 2% to 250psia.




      (2)  Reagents - The absorbing reagent was prepared with reagent




grade chemicals - 13.61g potassium dihydrogen phosphate, 14. 2g of




anhydrous disodium hydrogen phosphate and 10. Og of potassium iodide,




diluted to 1.0 liter with distilled water which had been passed  through




two  ion exchange columns and a millipore filter.  Prior to  initial use,




each  new batch of absorbing reagent was aged for 24 hrs. in the  dark




and  at room temperature.




          Standard iodine solution, 0..05N, was prepared by  carefully




weighing 16. Og of potassium iodide and 3.173g of iodine and  diluted




to exactly 0.5 liter.

-------
     (3)  Procedure - Exactly 10 ml of the absorbing reagent was




pipetted into each of two tandem impingers, carefully shielding the




solution from light whenever possible.




          Ambient air was aspirated through the assembled train for




sampling periods of 15 to 30 minutes, at flow rates between 50 to




300 ml/min.  Immediately following sampling, the exposed absorbing




solution was transferred first to a graduated cylinder, then to a




1 inch diameter colorimeter cuvette.  At the lower flow rates (i.e.,




less than 300 ml/min)9 no appreciable evaporation was observed in




either impinger.




     (4)   Analysis - Absorbance was determined within 10 to 15 minutes




after sampling, with a Bausch and Lomb Spectronic 20 Spectrophoto-




meter, at 352 nm and with unexposed absorbing reagent as the




reference.




          Standardization was checked each day during the test, with




an 0.0025N iodine standard solution prepared by pipetting 5 ml of the




0.05N standard stock solution into a volumetric flask and diluting




to 100 ml with absorbing reagent.  A series of 0.2, 0.4, 0.6, and 0.9 ml




portions of the diluted standard iodine were pipetted into volumetric




flasks and diluted to 25 ml with absorbing reagent and the absorbance




of this known series was determined at 352 nm.

-------
          Calculations - Absorbance and calculated normality of the

standard iodine solution was plotted  (refer to Fig.B-6), and the

standardization factor, M, was determined from the normality of the

standard solution at absorbance 1.0.

          03  (ppm)  = Sample absorbance x M/V  (Ref. A-l)

               where;  M (standardizations factor) = 1.224 x 10 I
                       I = Intercept at absorbance 1.0 and
                       V = Volume of sample in liters

          The above calculations were executed, assuming no deviation

from the standard conditions o:_ 760 mm of mercury and 25ฐ c, and the

air sample volume was used directly as recorded.

          Then:  Volume = 24,47 liters
                 ly !„ = 24.47 yl 03 and
                 10 ml IN iodine = 5 x 103 ymole I2 = 1.224 x 10 yl 03

          Procedure efficiency - Varying the sampling time, not

exceeding a maximum of 30 minutes or a minimum of 15 minutes and

obtaining the absorbance reading at intervals of 10 to 15 minutes

after sampling, the error observed between consecutive samples was

negligible.   Variations did occur among successive batches of absorbing

reagent, which gave only 93% reproducibility.
Ref.  B-l  Saltzman, B, E.,  Determination of Oxidants (including Ozone):
          Neutral Buffered-Potassium Iodide Method, Public Health Service
          Publication No. 999-AP-ll.

-------
A.C. VOLTAGI
REGULATOR


^

                                                                         LAMP CONTROL
AMB. AIR
  INLET


SCRUBBER
V


'
t
MASS FLOW
METER


                                            A.C.
                                          VOLTMETER
             12 v.d.c.
              SUPPLY
                                              !SHUTTER
                                               CONTROL
                                                          APERTURE
                                                            ADJ.
CALIBRATION
   UNIT
 LAMP
SUPPLY
                                    MA
                                   METER
                                                                    IMPINGERS
                              j  SPECTROMETER i
                                                 TIMER
                                               (ELECTRONIC
                                                COUNTER)

-------
••••••••*•••••••••••*ป*ซ••*•••••••••••••••••••••••••••••••>•••••••••••ป
immmtmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmm'           tmvmmmmmmmmmmmmmmt                         ^mmmmmmmmmmmmmmmmmmmmmm
mmmf*mmm*tmmmmmmmmfummmmmmmmmmmmmm*mmm*mmmmmmmmm*mปmmmmmm*m:m{mmmmmmmmt*mfmmm*mmf*mmmmm*mm*mmmmmmmmm*mmmmmmmmmm.muwmummm
mmmmmmmmmmmmmmmmmmm*mmmmmmmmfm*mmmmmmmmmmmmmmmmmmmmmmmmmmm1mmmmmmmmmtmmmmmmmmmmm.mmmmmmmmmmmmmmmtmm*mmmmmmmmmmmm.*mmmmm
,wm-mmmmm.mmmmmmmmmmmm.mmmmm*m.nmm,m.m.m,mmmmmmmm.mnmm-mm.m.m.mmm.m.mmmmm.mmm.m.v mmmmmwmmmmmmmmmmmmnmmmmmmmmmmmmmmmmmmmmw^mmmmmmmmmmttttm
immmmปmmmmmmmmmmmmmmummmmmmmmmm*mmmmmmm*mmmmmmmmmmmmmmmmmm*mmmm immmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmm
immmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmm^mmmmmmmmmmmmmmmm immmmmmm•••ป•***ป•!••••••••*•••••••••*••••••••••••*•••••
                                          .mmmmmmmmmmmปmmmmปmmm.mmmmmmmmmmmmmmmmmmmmmm*m^mmmmmmmmmmmm.mmmmmmmt
immmmm.mmmmmmmmmmmmmmmmmmmmmmmmmm'mmtmmm.mfm.mmmmmmmmmmmmmmmm.mmmmi mmmmmmummmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmmtmmtmmmmm
• •••••••••••••••••••••••••••••••••••••••••••••••••••••••***• !••••••••••••••ป*•••••••ซป•ซ•ป••••••ป••ป••*••ปปป•B
'••••ป•••ป••••••••ซ••••••••••••*••••••••••••••ซ•••ซ••••••*•ซ•••ซ••*•••*•••
                                                                                                                            • B BBIIP* ••••p5ซBpppซป*ipappa BBC ••
                                                                                                                            •a*aBBBaP*f*aปfcaaBBB*f•••aaaBaai
                                                                                                                            •BaaBBBaaaia BaBBaapiaM*BซBPซppซ
                                                                                                                            ••BBBaaVBaaaaaiBBaaiapBaaaaBBBBB

                                                                                                                            i:::::::::::::::::::::::::::::::
                                                                                                                            i:::::::::::::::::::::::::::::::
                                                                                                                                             •••••••••*•••••••
                                                                                                                                             ••••••••••••••••
 mmm mmmmmmmmmmm" mm m'l                                 mmmt mmmm                                                  \mmmm' •*•*
 aซi*ซBB.* ••BBBBBaa.pBBBBaaซBap •••BB>BB*pIIIBBB*PซP-•••BBBPBPBBBMHPIBPK
 ia*BปkBBB*ซBBpaBp*B•BBB>aซBBapnaBaMBBBBpppapซBBBpr ajaBJBBBBBB BBBBpซปBipBBBBBB.B.*aB BBaaBBBaBa BBป*ปBBBBซ * 4iwssปซซ*Bซmt*ซ*!SfflปKBi
 IBBB ••••&•••••••••••*••••*•••••••••••••• •••••paBBJB.                     1 *•*••••••• •••BBBPMBaPBMl                  EBBBPBBBBB
                iBaRaBa.BซaaaB*BaaaaBBaaBBBBBBaBBB*a••••••ป••••••••••••••••••••••••••••••••ammmmmmmmm'BซaaBBpa.NpBaป*PBBซap
 • ••   PBPPPBB BBaaaaBaaaBaaBBBaBaai                        BE      • ป™^ป-ป ปPBB-ซ--ซ aaBHuaavaaBa  ซpprjap*pii*pi   ••*mmr*—*m~
 mmm   BBBB?   aaapaaaaBBBBBBaBBBBBBBMHaBBBNBPBB'IBBBBBBBI *f   jซaปii i i at i *••••. Ji a*IBBBBBBBBB  BSI;        ซ•* j   a IBBB •> i i a
 BBB   BBBBaiBB BBBBBBBBBaBBaBBBBBBBBBBBaaBaBB*aB*BaBBBaBBBfc   apซhik*MavbBtfaBh'ซB*4llซ*B •BBBB'BB  VBpaBatfaaWBBbB   •BBB**B*BB
 ปBB   WBBBBB BVWBVBVBaBB'tfB'BaBBBBaaBBBBBB.aB'tfB'BBBBJVBaaaiBB1   ••ป•ป•••• BaBBBaaflBJB BBBBB HBaBH BB  BB^BBBtfBBBBItBBB   ••BBBVBBBB
 • Bi   PBPBBBP ••••••••••••••BB>iซBซ*B.B•••••••< ••!•ซ••••••   •••••••••••*•••••••••••••••aซBB  B', BB•Pซ••P•B•PB   ••••••••••
 ^ssป   isssR*ซ* BBBซaaaซปaซa*ซซaป|ซPBBBซp•••••ซป *sซ8itซsatt3Kw^   •ซ••••ซ••ป•••••••• ••••aMBซaBK  asnปttปS!*ซซปii*ปa   siซBซ*wปaeซ
 • •i   Bซ*Bปซ mmmmmmtmmmmmmmmmmmmmmmmmmmmmmmmwmmmmmfmmmmmm   mmmmmmtmmmmmmmmmmmmmmmmmimmmmmm*  immmmmmmmmmmmmm   BBBBBBBBBB
 •••   •ซซ••ป HBB*ซซBB>ซปซปซซmปmซBซซซaiB,*Bซiซa*' ซ••ซ•ซ•ซ•••ซ   ••iซBซปซปปB.ซmfcป*ซปซ*ซปซB,ซ,9           ••BaaaNVVBa   *•••••ซ*••
 •BaaliaBBBBBBBBaaBtfBBBBBBBBaaiBaBBBBBBaBaBBBBf ซBBBBBaaaBBBBBBBBBBjaBBaiBBBai*BBaiaiBPBBa;HIBBjBaBir"BBBBB1IIBBBBH*BBHaBBBBBBBVB*aBBl
 • •PBP>BPPpaBPซปซปP*ซBBBBflaBBPB*ซPBPBBP•••• I •••ปBaBa******•••••••••••••'••••••I*••••••• i . •1BBBBBaBซซBBBBBBaBa•••••BBBBaI
 BaaBB*BBBBBBBBBaBBaaaa.aaaBซBBBBBซซ9Bซ*BซBBBซBBBiaaBBBBBBBBaBaBซBBBซBi**B*BBaBBBBBBBBBB'.*>. ^••Bปซปปซป BB.BB* a'BBlBB •BBBBBBปBซ I
 • •••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••PMftrflWWHMBBVvPailBBHBMVaBWaWflBBVW
 •••••••••••••••••••••••••••••••••••••••I•••••••••••••ซ•ซ••••••••••ง•••••ซ••••••••^•••••B•••••••••••••••••••!•••*••••
 •••••• ••••••••••••ซ!• ••••••••••••••••••••• !•••••••• •••ftVMBfl*M ••••••*••• PBKVBWHaBVP'jMVBBVBM ••••••••*•! ••VliftBAflVH *•*••••*
 ••••••••*•••••ซ>ซ!•ป*•*••ซ••!••••••••••••••!           ป•••••••••••••ง•*••••••ปป•'••••••*^V
 ••••••••••••*•*!           !••••••••ซ•••I••••••••••••••••••••••••••••••••••••••••'•••••Pซซ•••••••••••••••••*•••••••ซ••
 • ••••••• ••••••••••••••••••••••*••••••• !••••ซ•••ซ ••••••••**••••••*••.••••••••••••.•••••••••••ฃ
 • •••••••••••••••••••••••••••••••••••a* !                  FซH••••••••ซ•••••••••• !•••••••••!
 • •••••••••••••••••••••••••!•nmmmmmmmmi •••••••••••B           ••••••••••••••••r••••••••••••••••••••••••••*ป•••••••••••
              •   ••       •••••• *•• i         •         ••     ••    •••ป•     ••ซ    ป   •   ป•••   ••*•     •••••   ••  ••
              •   i>       ••••••  •• i        tf         aซt     *s    awsww     ปป•    ป   ^   ปปปH   aปซ*     •••     ••  ••
                  •         ••*••  •• •        ซ         •*!?     ซป    ซWปKiB     t*ป*    •   *   •••    ••••     •••     MM  ••

                  :         :::::                        ::     ::..  :::       ::     :       :::    ::::     :::     ::  s:
                  i         •••••                        ••     ••*•  ••        ••     •       ป*•    ป•••     •••     ••  "•
                  •         •••••                        ปS     Iff ฃ  4iซ        *S     If       i>i*K    ซ>)ป      WWป     9^t  IS
                  •         •••••                        ••     ••ป I  ••        •*             8ฃซ    Ma*      ซปn     sซ  ซป
                  9         osซปs                        ซ•     ••>•!ง  ••        ••     •       •••    ••        ••*     ••  mm
                  5         •••••                        *•     mmn  ปr        ••     •       ป•*    an        *iปs>     *•  ••
                  •         •••••                        ••     ••    •'.        ซ•     •       •ซ•    •ป        ปซ•     ••  •*
                  •         ••ซ••                        mm     *      f --S        rt*     ป       ซปซ    sass        ซwป     iซs  w*
           !••     ป       •••••>•ซ              •       •••    • .      ••*     ปt      u     **ซsa    ;iซ          a    ••••
           ••    •••      •••••! ••         *     ••      •••   ' '*      *••     •••     •     •••••    ••      •••••    ••  ••
  •  ป     BBB   •••      BB*  SBBKBBBBBB  fmm   BBBBBIBBBB   mm    PP*BPPBPBP   *aa  mm   mmm.  BBBBBBBBBBBBBBBBB*    mmm
  m  mm     mm    mmm      BB>I  IpBilBpaaaB  mmm   aaaaaa*       ซป    •ป••••• BBP   ป*ป       ปซ   ซซ*ซ    ซซ;       ปsป    BBB
     B             *       mm i  mmmmmm  mm       Bป                 aซe9neซ*wซ    aw       •*   Bปa     >B       BBB    aa
                  ซ       •> s  *eB(iปffi  ซซ       ป*                       i*aซP    BB       BP   •*      BB       *B     a*
                  Ha       B'lflBaanVBBB       aa                 VBBBBBPBB    mm       mm   mm      mm       mm     mm
                          mim  liaiaa  pa       aa                 •• BBalaaa    BP       B*   BB      ปa       BB     •*
                          rซa  SaซซBB  BB       a*                 pa mmmmmmm    mm       mm   mm      mm       mm     mm
                          >iBBBBB.*BBBB       *B                 BBBBBaB*    BB       BP   •*      BB       BB     mm
                          IBB  ••••••  aa       aa                 *   BBliaaa    wป       ปซ   ซM      ซs       nซ     *ป
                          PPP  IaaซBB  BB       ปซ                 a   JsS)*ซBaซ    ซป       ซf>   ปn      i;a       a*;     w&
                          aai  iapBซa  BB       ซn                       ป•••    aa       BB    •      aa        m     mm
                          mmm  B.BBBBB  *a       aa                 a   aaaBaaa    BB       aa    ป       B        a     BB
                          we*  ซป*3ปซ  KB       BB                 a   BBaaaaa    aa       aa    •       a        a     BB
                          BซB  ipซซซa  BP       mr                 m   PBMซPBB    ซa       mm    m       m        m     mm
                          mmm  laaaaB  BB       m'                 e*iiปปซaottB    as       B*    ซ       ซ        *     *e
                          aa>  iaartaP  P        '.a                 eaซ4ปซaปซ    OFS       nn    n       ซ        ป     ia
                          BBB  SaaiBPB  •        *•                 B  BBBBBBBB    BB       BB    a       •        B     BB
                          BBBBBaaBBB  B        BB     B    BB    VB  BBBBB'BB    BB     B ••    B       A        m     BB
                          aaaaaaaiM  a        BB     m    mm    mm  aaBBaiaa    aa     B ••    B       •        a     BB
                          KซซMamffซ3  is       *isป    aa    ;ta           )ซaซa    aa     BBB    a      aa       aa     ••
           !a     •     aaaBBBaaaBBa  a       mmm   mmm    mm    mm  BBatBiaa    BB     a BB    a      BB   B   mmm     mm
           m     m     BBBBBBCBBBBa  •       BBB   BBB    ••    BB  ซBaBซBBB    BB     B •• BB B      BB   ง   BBB     ••
 ••  •  BBHB   mmmm   ป*•ป•,ซ*•ซ** Bซ     *a*aa   mmm   mmmm  BiซซซaซaBปaaB    a*     BBBB BB a    ••*   a   *aa   ซaซa
       Smmmmmm   •••••  •••  ••••••*••     •  •  •••    ••••*•••*•••••••••            ปป      am   &",              aa
       ซtsufซu   w*i tic;  ana  stปปซiซaซซst     n  ซ  ^MH    Sปsi5s*ซปa9  •••••••ซ            ae      aฅ   aป              iia
       ••••••   ••<•*  •••  mmmmfmmmm     *  •  •ซ•    •••••••••*  ••   ••••••            ••      ••   ••   ••••••ซ    ••
 •••••••<••••••ป••••**•••••••••••••••••••••••••••••••••••••*••                                                   Bซi?ป*wปnป'
 iปm.m.ปปtm.ปmmปปB. am*•••%*ปปปmmซป•ปป•ปปปปปปปปปป•••••• ซปปปปป*ปป• ปซซปปปป*ซป*ปปป***ปปปป*****ป******Pซปปซ••••ซ•• ซmซmmmปmiป*ปปmซ<
 <•••••>.••••••-.••••••••••••••••••••••*•••••••••••••••••••••••••••••••••ป•*••••ป••••••••••••••ฃ••••ป••••ป•ป•••••ป•••ซ•••

-------
    B-12. Chemiluminescent Disk Preparation

          The chemiluminescent disks are prepared by absorbing an

organic dye, Rhodamine B on a finely divided silicon-gel surface

which has been made moisture-insensitive by treatment with a hydrophobic

agent.  The high surface area material used as the substrate was

Eastman thin layer chromatographic-grade silica-gel sheet, Type

ECS-6061.  The dye compound was Rhodarodne-B.O (the hydrochloride

salt form), and the hydrophobic coating material was General Electric

silicone resin, type Sr-82.

     The recommended procedure for preparation of the chemiluminescent

disk surface is as follows:

     1.  Cut 2 inch diameter disks from a silica-gel sheet and dry

in an oven for 1 hour at 110 to 120ฐC.

     2.  Prepare a 10% (W/W) solution by diluting Sr-82 (60% resin)

1:6 with benzene.  Immediately after drying, dip the disk in the

hydrophobic resin solution for 1 minute.  Remove from the solution

and drain off the excess.  Place disks into oven and dry for 2A-48

hours at 110-120ฐC.

     3.  After  removing dried  disk  from oven, immediately  dip  the  disk

for 1 minute in a solution containing 0.1 gram per liter of Rhodamine-

BS0 in acetone.  Drai.n off the excess solution and return disk to

oven for about 5 minutes to evaporate off the remaining acetone.  Disks

prepared in this way are insensitive to moisture but should be protected

from the atmosphere, and bright light since the dye compound may

undergo slow oxidation.  The best practice is to store them in a

desiccator.
   Hodgeson, J. A, Krost, K. J., O'Keeffe, A. E, and Stevens, R. K.,
   "Chemiluminescent Measurements of Atmospheric Ozone," Anal. Chem.,

-------
                               APPENDIX C
                               PARTS LIST

     A parts list for each of the major sub-assemblies is given in this
appendix.  The non-standard parts are indicated by an RTI drawing number.
All other components are identified by manufacture and manufacturers'
part number.  Most of the parts listed are self-explanatory.

     The high voltage power supply (Power Design, Model 2K-10) , the Hh
15 Vdc supply (HP Model 60155C) , and the 12 Vdc supply (Deltron Model
C12-2.8) are not included since they are fully described by their res-
pective instruction manuals, which are included in the sub-systems
manuals.

-------
     C.I  Detector Assembly
Part No.  Description
001       Housing
010       PMT Housing
020       Photomultiplier
030       High Voltage Power
          Supply
040       Quartz Window
          (2% in. diam. x
          1/16 in. thick)
050       Jack
Manuf act lifer
RTI
Pacific Photometric

EMI
Power Designs Inc.
Mfg. Part No.
Dwg. No. 001
62 (modified
    length)
9558C
2K-10
Thermal American       Vitreosil
Fuzed Quartz Company

Precision Scientific   Little Jack
Co.
     C.2  Calibration Assembly
Part No.  Description
 100       Housing
 120       UV Lamp
 130       Lamp  Power  Supply
 140       Auto-transformer
 141       Knob
 150       Milliammeter
 160       Shutter  Solenoid
 170       Iris  Diaphragm
 180       Micrometer  Control
 190       Quartz tube
          (standard normal wall
          10.2mm o.d. x  8" L)
Manufacturer
 RTI
 Pen-Ray
 Pen-Ray
 Staco
 Raytheon
 Triplett
 Ledex
 Edmund Scientific
 Starrett
 Thermal-American
 Fuzed Quartz  Company
Mfg. Part No.
Dwg. No.  100
11SC-1B
SCT-3
121
70-4-26
430
H-1079-032
681
440-3RL

Spectrosil

-------
C.3  Plumbing Assembly
Part No.
1
2
3
4
5
6
7
8
9
10
11
12
C.4
Part No.
900
T-l
TP-1
T-2
5901
Kl
K2
K3
Bl
B2
PWR1
Description
Bulkhead Adapter
Male Elbow
All Tube Tee
Valve, fine
metering
Valve, fine
metering
Male Connector
Tie connector,
teflon
Reducing Union
Solenoid Valve
Flow Meter
Elbow (Teflon)
Pump
Timing Control
Description
Chassis
Cycle Timer
Cycle Programmer
Purge Timer
Mode Switch
Valve Relay
Valve Relay
Timer Relay
Lamps (Mode Switch)
Lamps (Shutter & Power)
+ 12 VDC Power Supply
Manufacturer
Swage lok
Swag el ok
Swag el ok
Nupro
Nupro
Swagelok
Chemplast
Chemplast
Valcor Engineering
Brooks Instrument
Division
Chemplast
Neptune Products, Inc.
Manufacturer
Bud Radio
Sealectroswitch
Corp.
11
Industrial Time
Corp.
Switch craft
Potter & Brumfield
Potter & Brumfield
Potter & Brumfield
GE
GE
Deltron
Mfg. Part No.
400-A1-4
400-2-2
40-3
B-4MA
B-4M
400-1-4
T-T0404-F
T-S0604-F
51C70HT34-6A
1555-OOC1A
T-L0404-F
4K
Mfg. Part No.
AC- 40 3
092-0166-500
092-0765-501
CM-10/10A24
37061/H83P/83P
KA11DG
KA11DG
KA11DG
330
328
C12-2.8

-------
C.5  Amplifier Assembly
Part No.
500
Rl
R2
R7
R8
R9
RIO
Rll
R12
R13
R14
R20 & 21
R22
R23 & 24
R25 & 26
R27
R28
R29
R27
R30
R31
R32
Cl
C2
C3
S501
S502
Al
A2
Ml

PWR1
Description
Chassis
100K Potentiometer
50K "
10K Potentiometer
10K
9 meg 1% Carbon
900K 1% Metal Film
100K " " "
IK "
99K "
qv- a ii ii
Manufacturer
RTI
Helipot
Helipot
Bourns
Bourns
IRC
IRC
IRC
IRC
IRC
IRC
[Trim resistors furnished with AD 183K
10K 10-Turn Pot.
1 meg "
200K " "
1000 Meg ohm
10K Trimpot
10K Potentiometer 10T
100 ohms 1% Metal Film
15K
13K
IK Potentiometer
0.9 yf
1.9 yf
4.1 yf
Sensitivity Switch
Time Constant Switch
Operational Amp.
Operational Amp.
Output meter
Knobs
Power Supply
Bourns
Bourns
Bourns
Victoreen
Beckman
Bourns
IRC
ii
ii
Helipot
Sprague
Sprague
Sprague
Centralab
Centralab
Analog Devices
Analog Devices
Simpson
Raytheon
Analog Devices
Mfg. Part No.
Dwg. No. 500
76PR 100K
76PR50K
3600S
300 7P






amplifiers]
3600S



76PR10K
3600S



76PR1K
151D-904X9020W2
15 ID- 19 5X90 35X2
151D-415X9020X2
P-505/(3)PS-21
P-503/(1)PS-21
311K
184J
1150
50-4-1G
AD902

-------
C.6
Part No.
500
S/H1.2&3
Al, 2&3
PWR3
Rl , 2 , 4 , 5
6,7,8&9
R3.10&11
C1,C2,C3
C4
M/D1
C.7
Part No.
701
800
801
802
803
804
805
806
401
301
Signal Conditioning Unit
Description
Chassis
Sample & Hold
Amplifier
Operational Amplifier
+ 15 V Power Supply
10 K Metal Film, 1%
1/4 W
100 K Metal Film, 1%
1/4 W
1 yf, 100V, 1%
.10 nf
Multiplier-Divider
Chassis and Cabinet
Description
Chassis
Cabinet
Front Panel
Rear Panel
Rear Door
Handles
Chassis Slides
Fan
Solenoid Valve Bracket
Clean-up Tower
Parts List
Manufacturer
RTI
Burr Brown
Analog Devices
Hewlett-Packard
IRC
IRC
Electronic Associate's,
Inc.
Sprague
Burr"Brown
Manufacturer
RTI
Optima
Optima
Optima
Optima
Optima
Chassis-Trak
Rotron Mfg. Co.
RTI
RTI
                                                             Mfg.  Part  No.

                                                             Dwg.  No.  500

                                                             BB4013/25


                                                             AD183J

                                                             60155C
                                                             113D-106C701500

                                                             BB4030/25
200
Bracket

Pump Mounting
Bracket
RTI
Mfg. Part No.

Dwg. No. 700

E-171920H

P-14

P-7

D-10

H-14

C-230-S-18

Muffin Fan

Dwgk No. 400

Dwg. No. 300


Dwg. No. 200

-------
      C. 8  Interlock Circuit
Part No.  Description
TR1       Transformer
          (12.6 VAC)
S101      Full Open Switch
S102      Full Closed Switch
S103      Door Open Switch
BZ1       Buzzer
B101      Full Open Lamp
B102      Full Closed Lamp
                        'Manufacturer
                         Chicago-Stancor

                         Robertshaw
                         Robertshaw
                         Robertshaw
                         Potter & Brumfield
                         Dialco/GE
                         Dialco/GE
                       Mfg. Part No.
                       P8130

                       1MD1-1A
                       1MD1-1A
                       1MD1-1A
                       12VDC
                       182-8430/327
                       182-8430/327
     C.9  AC Power
Part No.
SI
Fl
PS1
Description
Power Switch
Fuse
Power Strip
Manufacturer
Arrow Hart
Littlefuse
CBC Electronics
     Part No.
81024GB
3AG-10AMP
MO-G-D

-------
                              APPENDIX D

                           SUBSYSTEM MANUALS
     The manufacturer-supplied manuals or data sheets on certain major
components are included in a separate binder.  Reference should be made
to these as required for operation, maintenance and repair information.
The manuals/data sheets included are as follows:

     I.   DETECTOR UNIT

          A.  Photomultiplier (EMI 9558C)
          B.  HV Power Supply (Power Designs 2K-10)

    II.   AMPLIFIER AND SIGNAL CONDITIONER

          A.  Operational Amplifier (AD 183J) or (AD 184J)
          B.  Operational Amplifier (AD 311)
          C.  Sample/Hold Modules (BB 4034/25)
          D.  Multiplier/Divider Modules (BB 4030/25)
          E.  Operational Amplifier Power Supply (AD 902)
          F.  DC Power Supply (HP 60155C)
   III.    TIMING AND CONTROL

          A.   Programming Switch (Sealectrowsitch SSC-3A)
          B.   Programming Switch (Sealectroswitch SSC 11)
          C.   DC Power Supply (Deltron C-12-2.8)

    IV.    CALIBRATION VMT

          A.   UV Lamp (Pen-Ray 11SC-1C)
          B.   Lamp Power Supply (Pen-Ray SCT-1)
          C.   Lamp Control (Staco 100 BV)

     V.    PLUMBING ASSEMBLY

          A.   Teflon Fittings (CHEMPLAST)
          B.   Metering Valves (Nupro B4MA)
          C.   Valve Vernier (Nupro)
          D.   Pump (Neptune 4K)
          E.   Rotameter (Brooks 1550)
          F.   Drying Tower (Drierite)
          G.   Solenoid Valve (Valcor)

    VI.    CHASSIS AND CABINET ASSEMBLY

          A.   Exhaust Fan (Rotron Mark 4)

-------
                              APPENDIX E
                         •ENGINEERING DRAWINGS

     A complete set of engineering drawings for the machine shop
fabricated parts is included in a separate manual.   A listing of the
specific drawing sets are given below.
Detector Assembly
Calibration Assembly
Pump Mount
Valve Bracket
Tower Mount
Amplifier Chassis
Main Chassis
Panels
Drawing Set
Drawing Set
Drawing Set
Drawing Set
Drawing Set
Drawing Set
Drawing Set
Drawing Set
# 001
# 100
# 200
# 400
# 300
# 500
// 700
# 600

-------