USEPA CONTRACT LABORATORY PROGRAM


        STATEMENT OF WORK  •

               FOR  '

           ANALYSIS OF
    LOW CONCENTRATION ORGANIC

             OLC03.2
          December 2000

-------
                                                              540R005O6
                                  STATEMENT OF WORK

                                  TABLE OF CONTENTS



EXHIBIT A:    SUMMARY OF REQUIREMENTS

EXHIBIT B:    REPORTING AND DELIVERABLES REQUIREMENTS

EXHIBIT C:    TARGET COMPOUND LIST AND CONTRACT REQUIRED QUANTITATION LIMITS

EXHIBIT D:    ANALYTICAL METHODS

EXHIBIT E:    QUALITY ASSURANCE/QUALITY CONTROL PROCEDURES AND REQUIREMENTS

EXHIBIT F:    CHAIN-OF-CUSTODY,  DOCUMENT CONTROL,  AND WRITTEN STANDARD OPERATING
             PROCEDURES

EXHIBIT G:    GLOSSARY OF TERMS

EXHIBIT H:    DATA DICTIONARY AND FORMAT FOR DATA DELIVERABLES IN COMPUTER-
             READABLE FORMAT
                                                                             OLC03.2

-------
                           THIS PAGE  INTENTIONALLY LEFT BLANK
OLC03.2

-------
       EXHIBIT A





SUMMARY OF REQUIREMENTS
         A-l                                    OLC03.2

-------
                           THIS PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                    A-2

-------
                           Exhibit A - Summary of Requirements

                                    Table of Contents

Section                                                                              Page


1.0    PURPOSE	5

2.0    DESCRIPTION OF SERVICE 	 5

3.0    DATA USES  	5

4 .0    SUMMARY OF REQUIREMENTS	•	5
       4.1    Introduction to the  Organic Low Concentration
              Statement of Work	5
       4.2    Overview of Major  Task Areas   	6
       4.3    Technical and Management Capability	•	13
                                           A-3                                     OLC03.2

-------
                           THIS PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                    A-4

-------
                                                                Exhibit A -- Sections 1-4
                                                                                  Purpose
1.0    PURPOSE
       The purpose of the Low Concentration (Water)  Organic analytical service
       is to provide analytical data for use by the U.S. Environmental
       Protection Agency (USEPA)  in support of the investigation and clean-up
       activities under the Comprehensive Environmental Response, Compensation,
       and Liability Act of 1980 (CERCLA) and the Superfund Amendments and
       Reauthorization Act of 1986 (SARA).   Other USEPA Program Offices that
       have similar analytical data needs also use this service.

2.0    DESCRIPTION OF SERVICE

       The organic analytical service provides a contractual framework for
       laboratories to apply USEPA Contract Laboratory Program  (CLP) analytical
       methods for the isolation, detection, and quantitative measurement of 50
       volatile,  65 semivolatile, and 28 pesticide/Aroclor target compounds in
       water samples.  The analytical service provides the methods to be used
       and the specific contractual requirements by which USEPA will evaluate
       the data.   This service uses Gas Chromatography/Mass Spectrometry
       (GC/MS) and Gas Chromatography/Electron Capture Detector  (GC/ECD)
       methods to analyze the target compounds.

3.0    DATA USES

       This analytical service provides data which USEPA uses for a variety of
       purposes such as: determining the nature and extent of contamination at
       a hazardous waste site; assessing priorities for response based on risks
       to human health and the environment; determining appropriate clean-up
       actions; and determining when remedial actions are complete.  The data
       may be used in all stages in the investigation of hazardous waste sites,
       including: site inspections; Hazard Ranking System (HRS) scoring;
       remedial investigation/feasibility studies; remedial design;
       treatability studies; and removal actions.  In addition, this service
       provides data that are available for use in Superfund enforcement/
       litigation activities.

4.0    SUMMARY OF REQUIREMENTS

4.1    Introduction to the Organic Low Concentration Statement of Work

       This Statement of Work  (SOW) is designed as part of the documentation
       for a contract between USEPA and a commercial laboratory performing
       analyses in support of USEPA Superfund programs.  The SOW is comprised
       of eight exhibits.  Exhibit A provides an overview of the SOW and its
       general requirements.  Exhibit B contains a description of the reporting
      .and deliverables requirements, in addition to the data reporting forms
       and instructions.  Exhibit C specifies the Organic Target Compound list
       for this SOW with the contract required quantitation limits for the
       sample matrix.  Exhibit D details the required analytical procedures to
       be used with this SOW and resulting contracts.  Exhibit E provides
       descriptions of required Quality Assurance/Quality Control  (QA/QC),
       Standard Operating Procedures (SOPs), QA/QC performance, and the
       reporting of data.  Exhibit F contains Chain-of-Custody and sample
       documentation requirements which the Contractor shall follow.  To ensure
       proper understanding of the terms utilized in this SOW, a glossary can

                                           A-5                                    OLC03.2

-------
Exhibit A -- Section 4
Summary of Requirements (Con't)

       be found in Exhibit G.  When a term is used in the text without
       explanation, the glossary meaning shall be applicable.  Specifications
       for reporting data in computer-readable format appear in Exhibit H.

4.2    Overview of Maj or Task Areas

       For each sample, the Contractor shall perform the tasks described in
       each section.  Specific requirements for each task are detailed in the
       exhibits referenced.

4.2.1     Task I:  Sample  Receiving,  Storage,  and Disposal

4.2.1.1       Chain-of-Custody

              The Contractor shall receive and maintain samples under proper
              Chain-of-Custody.  All associated document control and inventory
              procedures shall be developed and followed.  Documentation
              described herein shall be required to show that all procedures are
              strictly followed.  This documentation shall be reported as the
              Complete Sample Delivery Group File  (CSF)  (Exhibit B).   The
              Contractor shall establish and use appropriate procedures to
              handle confidential information received from USEPA.

4.2.1.2       Sample Scheduling/Shipments

              Sample shipments to the Contractor's facility will be scheduled
              and coordinated by the Contract Laboratory Program  (CLP) Sample
              Management Office  (SMO).  The Contractor shall communicate with
              SMO personnel by telephone,.as necessary throughout the process of
              sample scheduling, shipment, analysis, and data reporting, to
              ensure that  samples are properly processed.

4.2.1.2.1         Samples  will be shipped routinely to the  Contractor through an
                  overnight delivery service.   However,  as  necessary,  the
                  Contractor shall be responsible for any handling or processing
                  of the receipt of sample shipments.   This includes  the pick-up
                  of samples at the nearest servicing airport,  bus station,  or
                  other carrier within the Contractor's geographical  area.   The
                  Contractor shall be available to receive  sample shipments at
                  any time the delivery service is operating,  including
                  Saturdays.

4.2.1.2.2         If there are problems with the samples (e.g.,  mixed media,
                  containers broken or leaking)  or sample documentation and
                  paperwork [e.g.,  Traffic Reports (TRs)  not with shipment,
                  sample and TR do not correspond], the Contractor shall
                  immediately contact SMO for resolution.  The Contractor shall
                  immediately notify SMO regarding any problems and laboratory-
                  conditions that affect the timeliness of  analyses and data
                  reporting.  In particular,  the Contractor shall immediately
                  notify SMO personnel in advance regarding sample data that will
                  be delivered late and shall specify the estimated delivery
                  date.
 OLC03.2                                    A-6

-------
                                                                    Exhibit A --  Section  4
                                                          Summary  of  Requirements  (Con't)

4.2.1.2.3        To monitor the temperature of the sample shipping cooler more
                 effectively, each USEPA Regional Office may  include  a  sample
                 shipping cooler temperature blank with each  cooler shipped.
                 The temperature blank will be clearly labeled: USEPA COOLER
                 TEMPERATURE INDICATOR.  The Contractor shall record  the
                 presence or absence of the cooler temperature indicator bottle
                 on Form DC-1, Item 9.

4.2.1.2.3.1          When the USEPA Regional Office supplies a cooler temperature
                     indicator bottle in the sample shipping cooler, the
                     Contractor shall use the USEPA supplied cooler temperature
                     indicator bottle to determine the cooler temperature.   The
                     temperature of the cooler shall  be measured at the time of
                     sample receipt by the Contractor.

4.2.1.2.3.2          The temperature of the sample shipping cooler shall be
                     measured and recorded immediately upon opening the cooler,
                     and prior to unpacking the samples or removing the packing
                     material.

4.2.1.2.3.3          To determine the temperature of  the cooler, the Contractor
                     shall locate the cooler temperature indicator bottle in the
                     sample shipping cooler, remove the cap, and insert a
                     calibrated thermometer into the  cooler temperature indicator
                     bottle.   Prior to recording the  temperature,  the Contractor
                     shall allow a minimum of 3 minutes,  but not greater than 5
      ••«"             minutes,  for the thermometer to  equilibrate with the liquid
. .:r.  •'-<'-             in the bottle.  At a minimum, the calibrated thermometer
: •<•:•   •':*.'-•             (±1°C)  shall have a measurable range of 0-50°C.  Other
 •:•:>,   *'..             devices which can measure temperature may be used if they
  -;.:   Ki             can be calibrated to ±1°C and have a range of 0-50°C.   If a
                     temperature indicator bottle is  not present in the cooler,
                     an alternative means of determining cooler temperature shall
                     be used.   Under no circumstances shall a thermometer or any
                     other device be inserted into a sample bottle for the
                     purpose of determining cooler temperature.  The Contractor
                     shall contact SMO and inform them that a temperature
                     indicator bottle was not present in the cooler.  The
                     Contractor shall document the alternative technique used to
                     determine cooler temperature in the Sample Delivery Group
                     (SDG)  Narrative.

4.2.1.2.3.4          If the temperature of the sample shipping cooler's
                     temperature indicator exceeds 10°C,  the Contractor shall
                     contact SMO and inform them of the temperature deviation.
                     SMO will contact the Region from which the samples were
                     shipped for instruction on how to proceed.  The Region will
                     either require that no sample analysis(es) be performed or
                     that the Contractor proceed with the analysis(es).   SMO will
                     in turn notify the Contractor of the Region's decision.   The
                     Contractor shall document the Region's decision and the EPA
                     sample numbers of all samples for which temperatures
                     exceeded 10°C in the SDG Narrative.
                                           A-7                                     OLC03.2

-------
Exhibit A -- Section 4
Summary of Requirements  (Con't)

4.2.1.2.3.5          The Contractor shall record the temperature of the cooler on
                     Form DC-1, under Item 10 - Cooler Temperature, and in the
                     SDG Narrative.

4.2.1.2.4        The Contractor  shall accept all samples scheduled by  SMO,
                 provided that the  total number of  samples received  in any
                 calendar month  does not exceed the monthly limitation expressed
                 in the  contract.   Should the  Contractor elect  to  accept
                 additional  samples, the Contractor shall remain bound by all
                 contract requirements  for  analysis of  those samples accepted.

4.2.1.2.5        The Contractor  is  required to retain unused sample  volume  and
                 partially used  sample  volume  in original sample containers for
                 a period of 60  days after  data submission.  From  time of
                 receipt until analysis, the Contractor shall maintain all  water
                  (preserved  and  unpreserved) samples  at 4°C  (±2°C).

4.2.1.2.6        The Contractor  shall be required to  routinely  return  sample
                 shipping containers  (e.g.,  coolers)  to the appropriate sampling
                 office within 14 calendar  days following shipment receipt  (see
                 clause titled,  "Government Furnished Supplies  and Materials").

4.2.1.3       Modified Analysis

              The  Contractor may be requested by USEPA  to perform modified
              analyses.   These modifications will be within the scope  of  this
              SOW  and may include, but  are not limited  to, analysis  of
              additional analytes and/or lower quantitation limits.  These
              requests will  be made by.  the USEPA Regional. CLP Project  Officer
               (CLP PO),  USEPA  OERR Analytical  Operations/Data Quality  Center
               (AOC) Organic  Program Manager, and Contracting Officer (CO) in
              writing, prior to  sample  scheduling.  If  the Contractor
              voluntarily elects to perform these modified analyses, these
              analyses will  be performed with  no' increase in per sample price.
              All  contract requirements specified in  the SOW/Specifications will
              remain in  effect unless the USEPA CO  provides written  approval for
              the  modification(s) and a waiver for  associated defects.  The
              USEPA CO approval  must be obtained prior  to sample scheduling.

4.2.2     Task II: Sample Preparation and Analysis

4.2.2.1      Overview

              The  Contractor is  advised that the samples received  under this
              contract  are usually  from known  or suspected hazardous waste  sites
              and  may  contain  high  (greater than 15 percent) levels  of organic
              and  inorganic  materials of a  potentially  hazardous nature and of
              unknown  structure  and concentration,  and  should be handled
              throughout the analysis with  appropriate  caution.  It  is the
              Contractor's responsibility to take all necessary measures  to
              ensure laboratory  safety.

 4.2.2.2      Sample analyses  will  be scheduled by  groups of samples,  each
              defined  as a Case  and identified by a unique USEPA Case  number
              assigned by SMO.  A Case  signifies a  group of samples  collected at

 OLC03.2                                    A-8

-------
                                                                   Exhibit A -- Section 4
                                                          Summary of Requirements  (Con't)

              one site.or geographical area over a finite time period,  and will
              include one or more field samples with associated blanks.  Samples
              may be shipped to the Contractor in a single shipment or multiple
              shipments over a period of time,  depending on the size of the
              Case.

4.2.2.2.1        A Case consists of one or more SDGs.  An SDG  is defined by the
                 following, whichever is most  frequent:

                 •    Each Case of field samples received,  or

                 •    Each 20 field samples [excluding Performance Evaluation
                      (PE)  samples]  within a Case, or

                 •    Each 7 calendar day period during which field samples in a
                      Case are received (said period beginning with the receipt
                      of the first sample in the SDG).

                 •    In addition,  all samples and/or sample fractions assigned
                      to an SDG must have been scheduled under the same
                      contractual turnaround time.  Preliminary Results have no
                      impaqt on defining the SDG.

4.2.2.2.2        Samples shall be assigned to  SDGs at the time the samples are
                 received,  and shall not be assigned retroactively.  However, PE
      •"-'         samples received within a Case shall be assigned to an SDG
      -'••>•         containing field samples for  that Case.

4.2.2.2^3        Each sample  received by the Contractor will be labeled with an
      ip-         EPA sample number, and accompanied by a TR bearing the sample
      -•'-*'         number and descriptive information regarding  the sample.  The
      s-         Contractor shall complete and sign the TR, recording  the date
                 of  sample  receipt and sample  condition on receipt for each
                 sample container.

4.2.2.2.4        The Contractor  shall submit signed copies of  TRs for  all
                 samples in an SDG to SMO within  three working days following
                 receipt of the  last  sample in the SDG.  Faxed copies  of TRs do
                 not meet this requirement.  TRs  shall be submitted in SDG sets
                  (i.e., all TRs  for an SDG shall  be clipped together)  with an
                 SDG Cover  Sheet containing information regarding the  SDG, as
                 specified  in Exhibit B.

4.2.2.2.5        USEPA Case numbers,  SDG numbers, and EPA sample numbers shall
                 be  used by the  Contractor in  identifying samples received under
                 this contract,  both verbally  and in reports/correspondence.

4.2.2.3       If insufficient sample volume (less than the required amount)  is
              received to perform the analysis, the Contractor shall contact SMO
              to apprise them of the problem.   SMO will contact the Region for
              instructions.   The Region will either approve that no sample
              analysis be performed,  or require that a reduced volume be used
              for the sample analysis.  No other changes in the analysis will be
              permitted.  SMO will notify the Contractor of the Region's
                                           A-9                                     OLC03.2

-------
Exhibit A -- Section 4
Summary of Requirements (Con't)

              decision.  The  Contractor  shall  document  the Region's  decision in
              the  SDG Narrative.

4.2.2.4       Analytical Techniques:  The Target  Compounds listed in  Exhibit  C
              shall be identified,  as described  in the  methodologies given in
              Exhibit D.   Automated computer programs may be used to facilitate
              the  identification of compounds.

4.2.2.5       Preparation  Techniques. The Contractor will prepare samples as
              described in Exhibit  D. For semivolatile and pesticide/Aroclor
              samples, an  aliquot is extracted with a solvent  and concentrated.
              The  concentrated extract is subjected to  cleanup procedures and
              then analyzed by Gas  Chromatography/Mass  Spectrometry  (GC/MS)  for
              semivolatile or Gas Chromatography/Electron Capture Detector
              (GC/ECD) for the pesticide/Aroclor target compounds listed in
              Exhibit C.   For volatile samples,  an aliquot  is  purged with an
              inert gas,  trapped on a solid sorbent, and then  desorbed onto the
              GC/MS for analysis of the  target compounds listed in Exhibit C.

4.2.2.6       Qualitative Verification of Compounds.  The volatile and
              semivolatile compounds identified by GC/MS techniques  shall be
              verified by an analyst competent in the interpretation of mass
              spectra by comparison of the suspect mass spectrum to  the mass
              spectrum of a standard of  the suspected compound.  This procedure
              requires the use of multiple internal standards.

4.2.2.6.1         If  a compound  initially identified by GC/MS  techniques  cannot
                  be  verified, but  in the technical judgment of  the  mass  spectral
                  interpretation specialist  the  identification is  correct, then
                  the Contractor shall report  that  identification  and proceed
                  with quantitation.

4.2.2.6.2         The pesticide/Aroclor  compounds  identified by GC/ECD  techniques
                  shall be verified by an analyst  competent in the interpretation
                  of  gas chromatograms and by  comparison of the retention times
                  of  the suspected  unknowns  with the retention times of
                  respective  standards of the  suspected compounds.   Compounds
                  shall also  be  confirmed by GC/MS  techniques  if the compounds
                  are of sufficient concentration  to be detected by  the GC/MS.

4.2.2.7       Quantitation of Verified Compounds.  The  Contractor shall
              quantitate  components  identified by GC/MS techniques by the
              internal standard method stipulated in Exhibit D.  Where multiple
              internal standards are  required by USEPA, the Contractor shall
              perform quantitation utilizing the internal  standards specified  in
              Exhibit D.  The Contractor  shall quantitate  components analyzed by
              GC/ECD techniques by the external standard method stipulated  in
              Exhibit D.  The Contractor  shall also perform an  initial
              calibration, verify  its linearity, determine the breakdown of
              labile components, and determine calibration factors  for all
              standards analyzed by  GC/ECD  techniques  as described  in Exhibit D.

4.2.2.8       Tentative Identification of Non-Target Sample Components.  For
              each analysis  of  a sample,  the Contractor shall conduct mass
              spectral library  searches  to  determine tentative  compound

OLC03.2                   .                A-10

-------
                                                                   Exhibit A -- Section 4
                                                          Summary of Requirements (Con't)

              identifications  as  follows:  For each volatile sample,  the
              Contractor shall conduct a search to determine the possible
              identity of up to 30  organic compounds  of  greatest concentration
              which are not deuterated monitoring compounds or internal
              standards and are not listed in Exhibit C  under volatiles or
              semivolatiles.   For each semivolatile sample,  the Contractor shall
              conduct  a search to determine  the possible identification of up to
              30  organic compounds  of greatest  concentration which are not
              surrogates or internal standards  and are not  listed in Exhibit  C
              under volatiles  or  semivolatiles.  In performing searches,  the
              NIST/EPA/NIH (May 1992 release or later) and/or Wiley (1991
              release  or later),  or equivalent,  mass  spectral library shall be
              used.

              NOTE:  Substances with responses less than  10  percent of the
              nearest  internal standard are  not required to be searched in this
              fashion.

4.2.2.9       Quality  Assurance/Quality Control (QA/QC)  Procedures

4.2.2.9.1        The Contractor shall strictly adhere to all specific QA/QC
                 procedures prescribed in Exhibits D and E.  Records documenting
                 the use of the protocol shall be maintained in accordance with
                 the document control procedures prescribed  in Exhibit F, and
                 shall be reported in accordance with Exhibits B and H.

4.2.2F9-. 2        The Contractor shall maintain a Quality Assurance Plan  (QAP)
     :-'-          with the objective of providing sound analytical chemical
     •**'••          measurements.  This program shall incorporate the QC
      ~          procedures,  any necessary corrective action, and all
                 documentation required during data collection, as well as the
                 quality assessment measures performed by management to ensure
                 acceptable data production.

4.2.2.9.3        Additional QC shall be conducted in the form of the analysis of
                 Performance Evaluation samples submitted to the laboratory by
                 USEPA.  Unacceptable results of all such QC or Performance
                 Evaluation samples may be used as the basis for an equitable
                 adjustment to reflect the reduced value of the data to USEPA or
                 rejection of the data for specific analyte(s) within an SDG or
                 the entire SDG.   Also,  unacceptable results may be used as the
                 basis for contract action.   "Compliant performance" is defined
                 as that which yields correct analyte identification and
                 concentration values,  as determined by USEPA, as well as
                 meeting the contract requirements for analysis  (Exhibit D),
                 QA/QC  (Exhibit E), data reporting and other deliverables
                 (Exhibits B and H), and sample custody, sample documentation,
                 and Standard Operating Procedure (SOP)  documentation  (Exhibit
                 F).  As an alternative to data rejection,  USEPA may require re-
                 analysis of non-compliant samples.   Re-analysis will be
                 performed by the Contractor at no additional cost to USEPA,
                 unless it is determined that the Performance Evaluation
                 sample(s) was defective.
                                          A-ll                                    OLC03.2

-------
Exhibit A -- Section 4
Summary of Requirements (Con't)

4.2.3     Task III: Sample Reporting and Resubmission of Data

4.2.3.1       USEPA has provided,  to the Contractor,  formats  for the reporting
              of  data  (Exhibits  B  and H).   The Contractor shall  be responsible
              for completing and submitting analysis  data sheets and computer-
              readable data on diskette (or via an alternate  means of electronic
              transmission approved in advance by USEPA)  in a format specified
              in  this  SOW and within the time specified in Exhibit B,  Section
              1.1.

4.2.3.2       Use of formats other than those designated by USEPA will be deemed
              as  non-compliant.  Such data are unacceptable.   Resubmission in
              the specified format, at no additional  cost to  USEPA, shall be
              required.

4.2.3.3       Computer-generated forms may be submitted in the hardcopy Sample
              Data Package(s), provided that the forms are in exact USEPA
              format.   This means  that the order of data elements is the same as
              on each USEPA-required form,  including form numbers and titles,
              page numbers,  and header information.

4.2.3.4       If  the submitted data package does not conform to the specified
              contractual or technical criteria, the Contractor will be required
              to resubmit the data package with all deficiencies corrected at
              its own expense.  The Contractor will respond within seven days to
              requests for additional information or explanations that result
              from the Government's inspection activities.  If the Contractor is
              required to submit or resubmit data as a result of a Regional
              request, the data shall be clearly marked as ADDITIONAL DATA.  The
              Contractor shall include a cover letter which describes which data
              are-being delivered, to which USEPA project the data pertain, and
              who requested the data.  Any and all resubmissions must be in
              accordance with the documentation requirements of this SOW.

4.2.3.5       The data reported by the Contractor on the hardcopy data forms and
              the associated computer-readable data submitted by the Contractor
              on diskette  (or via an alternate means of electronic transmission,
              if approved in advance by USEPA) shall contain identical
              information.  If discrepancies are found during Government
              inspection, the Contractor shall be required to resubmit either
              the corrected hardcopy forms or the corrected computer-readable
              data, or both sets of corrected data, at no additional cost  to
              USEPA.

4.2.3.6       In  addition, the Contractor must be aware of the  importance  of
              maintaining the integrity of the data generated under this
              contract, since it is used to make major decisions  regarding
              public  health and environmental welfare.  The data may also  be
              used  in litigation against Potentially Responsible  Parties  (PRPs)
              in  the  enforcement of Superfund legislation.
 OLC03.2                                   A-12

-------
                                                                   Exhibit A -- Section 4
                                                          Summary of Requirements  (Con't)

4.3    Technical and Management Capability

4.3.1     Personnel

          The Contractor  shall  have  adequate personnel at all  times during the
          performance of  the  contract to ensure that USEPA  receives data that
          meet  the terms  and  conditions of the contract.

4.3.2     Instrumentation

          The Contractor  shall  have  a sufficient Gas Chromatograph/Electron
          Capture/Data System (GC/ECD/DS), Gas Chromatograph/Mass
          Spectrometer/Data System (GC/MS/DS), including magnetic tape  storage
          devices to meet all the  terms and conditions of the  contract.

4.3.3     Facilities.

          The Contractor  shall  maintain a facility  suitable for  the receipt,
          storage, analysis of  the samples, and delivery of the  product meeting
          the terms  and conditions of the contract.
                                          A-13
                                                                                   OLC03.2

-------
               EXHIBIT B
REPORTING AND DELIVERABLES REQUIREMENTS
                   B-l                                     OLC03.2

-------
                           THIS PAGE  INTENTIONALLY  LEFT BLANK
OLC03.2                                    B-2

-------
                    Exhibit  B  -  Reporting  and Deliverable  Requirements

                                    Table of Contents
Page
1.0    CONTRACT REPORTS/DELIVERABLES DISTRIBUTION 	  5
       1.1    Report Deliverable Schedule	5
       1.2    Distribution    	  8

2.0    REPORTING REQUIREMENTS AND ORDER OF DATA DELIVERABLES   .•	   10
       2.1    Introduction	   10
       2.2    Resubmission of Data   	   10
       2.3    Quality Assurance Plan (QAP)  and Standard
              Operating Procedures   	  	   11
       2.4    Sample Traffic Reports   	   11
       2.5    Sample Data Package	   12
       2.6    Complete SDG File	   25
       2.7    Data in Computer-Readable Format   	   26
       2.8    Preliminary Results  	   26
       2.9    GC/MS and GC/ECD Tapes   	   27
       2.10   Extracts   	   27

3.0    FORM INSTRUCTIONS	   28
       3.1    Introduction	   28
       3.2    General Information  	   28
       3.3    Header Information   ......  	   29
       3.4    Organic Analysis Data Sheet  (Form I, All Fractions)  	   33
       3.5    Organic Analysis Data Sheet: Tentatively Identified Compounds
               (Form I LCV-TIC and Form I LCSV-TIC)   	   36
       3.6    Deuterated Monitoring Compound  (DMC) Recovery
               (Form II LCV-1, LCV-2 and Form II LCSV-1, LCSV-2)	37
       3.7    Surrogate Recovery  (Form II LCP)   	   38
       3.8    Matrix Spike/Matrix Spike Duplicate Recovery
               (Form III, All Fractions, LCV, LCSV, LCP-1)	   39
       3.9    Method Blank Summary  (Form IV, All.Fractions)  	   41
       3.10   GC/MS Instrument Performance Check
               (Form V LCV and Form V LCSV)   	•	   42
       3.11   GC/MS Initial Calibration Data  (Form VI LCV-1, LCV-2,
              LCV-3 and Form VI LCSV-1, LCSV-2, LCSV-3)	43
       3.12   GC/EC Initial Calibration Data  (Form VI LCP-1, LCP-2, LCP-3)   	   43
       3.13   GC/MS Continuing Calibration Data  (Form VII LCV-1,
              LCV-2, LCV-3 and Form VII LCSV-1, LCSV-2, LCSV-3)  	  ...   46
       3.14   GC/ECD Calibration Verification Summary
               (Form VII, LCP-1, LCP-2)   	46
       3.15    Internal Standard Area and RT Summary
               (Form VIII LCV and Form VIII LCSV-1, LCSV-2)	   47
       3.16    Pesticide Analytical Sequence  (Form VIII LCP)  	   49
       3.17    Pesticide Cleanup Summary  (Form IX, LCP)   	   50
       3.18    Pesticide/Aroclor Identification  (Form X, LCP-1,  LCP-2)  	   51
       3.19    Sample Log-In Sheet  (Form DC-1)  .	   52
       3.20    Complete SDG File  (CSF)  Inventory Sheet  (Form DC-2)	   53

4.0    DATA REPORTING  FORMS	54
                                            B-3                                    OLC03.2

-------
                            THIS  PAGE  INTENTIONALLY LEFT BLANK
OLC03.2                                    B-4

-------
                                                                   Exhibit B -- Section 1
                                               Contract Reports/Deliverables Distribution

1.0    CONTRACT REPORTS/DELIVERABLES DISTRIBUTION

1.1    Report Deliverable Schedule.  The following table reiterates the
       contract reporting and deliverable requirements specified in the
       Contract Schedule (Performance/Delivery Schedule) and specifies the
       distribution that is required for each deliverable.  The turnaround
       times for items B through E listed below are 7, 14, and 21 days.

       NOTE: Specific recipient names and addresses are subject to change
       during the term of the contract.  The USEPA Analytical Operations/Data
       Quality Center (AOC) Organic Program Manager will notify the Contractor
       in writing of such changes when they occur.

                                         Table 1


                                                                       Distribution


                                                                         s
                               NO. of                                i   s*.
                                                                     CQ   fX
             Item              Copies*    Delivery Schedule


 A.2         Sample Traffic        1      3 working days            X
             Reports   '                   after receipt of
                                          last sample in
                                          Sample Delivery
                                          Group (SDG)-1

 B.2         Sample Data           1      XXB  days after            X
             Package0                     receipt of last
                                          sample in SDG.
                                                                                     N
 C.2         Data in               1      XXs  days after            X    X
             Computer-                    receipt of last
             Readable                     sample in SDG.
             Format

 D.2- 3       Complete SDG          1      XXB  days after                  X
             File                         receipt of last
                                          sample in SDG.

 E.5         Preliminary           1      Within 48 hours           X     X
             Results  (VOA                 after receipt of
             Analyses)                    each sample at
                                          laboratory, if                                 .
                                          requested.

             Preliminary           1      Within 72 hours           X     X
             Results  (SV                  after receipt of
             and Pest                     each sample at
             Analyses)                    laboratory, if
                                          requested.
                                            B-5                                    OLC03.2

-------
Exhibit B -- Section 1
Contract Reports/Deliverables Distribution  (Con't)
                                                                      Distribution
              Item
                   No. of
                   Copies
                                           Delivery
                                           Schedule
                                                                   o
                                                            §
                                                            -H
                                                            Ol
                                                            0)
                                                            PS
             Standard
             Operating
             Procedures--
             Technical and
             Evidentiary
                             Revise within  60
                             days  after
                             contract  award.

                             Submit within  7
                             days  of receipt
                             of  written
                             request to
                             recipients  as
                             directed.
                            As directed
             Quality
             Assurance  Plan
                             Revise within  60
                             days  after
                             contract  award.
             GC/MS  and
             GC/ECD Tapes.
                      Lot
Submit within 7
days of receipt
of written
request to
recipients as
directed.

Retain for 3
years after data
submission.
                                                                       As directed
                                                          As  directed
              Extracts
                                    Lot
                              Submit  within 7
                              days  after
                              receipt of
                              written request
                              by CLP  PO.

                              Retain  for 365
                              days  after data
                              submission.
                                                                       As directed
  J.e
Method
Detection Limit
Study
Submit within 7
days after
receipt of
written request
by CLP PO or
SMO, at USEPA's
direction.

Submit to USEPA
within 7 days
after receipt of
written request
by CLP PO or
SMO, at USEPA's
discretion.
                                                                       As directed
OLC03.2
                                            B-6

-------
                                                                   Exhibit B -- Section 1
                                       Contract Reports/Deliverables Distribution  (Con't)
Footnotes:
      AThe number of copies specified is the number of copies required  to be
delivered to each recipient.

      BThe number of days associated with these elements will be provided in
the associated laboratory contract document,  and will also be provided at the
time of the sample scheduling by the Sample Management Office (SMO)
Contractor.

      Contractor-concurrent delivery to USEPA designated  recipient (e.g.,
QATS) may be required upon request by the CLP PO.  Retain for 365 days after
data submission, and submit as directed within 7 days after receipt of written
request by the CLP PO.
      *A Sample  Delivery Group (SDG)  is  a group  of  samples  within a Case,
received over a period of 7 days or less,  not exceeding 20 samples,  (excluding
PE samples) and scheduled under the same contractual turnaround  (Preliminary
Results have no impact on defining the SDGs).   Data for all samples  in the SDG
are due concurrently.  The date of delivery of the SDG or any samples within
the SDG is the date that the  last sample in the SDG is received.  See Exhibit
A for further description.

      2DELIVERABLES ARE TO BE REPORTED TOTAL AND COMPLETE.   Concurrent delivery
required.  Delivery shall be  made such that all designated recipients receive
the item on the same calendar day.  This includes resubmission of both the
hardcopy and electronic deliverable.  The date of delivery of the SDG, or any
sample within the  SDG,  is the date all samples have been delivered.  If the
deliverables are due on a Saturday, Sunday, or Federal holiday,  then they
shall be delivered on  the next business day.  Deliverables delivered after
this time will be  considered  late.

      3Complete  SDG File will contain the  original  Sample Data Package plus all
of the original documents described under Section 2.6.

      4See Exhibit E and Exhibit F for a more  detailed description.

      5If requested at the time of sample  scheduling,  the Contractor shall
provide  Preliminary Results,  consisting of  Form I and  Form I  TIC analytical
results, by fraction,  for  field  and QC  sample analyses via  telefacsimile
 (fax).   The Contractor may submit Preliminary Results  in electronic format
after obtaining permission from  USEPA.  The Contractor will be notified of the
fax number or E-mail  address  at  the time of sample scheduling.   Sample Traffic
Reports  (TRs) and SDG cover  sheets shall be submitted  with the Preliminary
Results.  The Contractor shall  contact  SMO  after confirming transmission.  The
Contractor shall  document  all communication in  a telephone contact  log.

      6Method Detection Limit Study is performed annually or for each new
instrument, whichever is more frequent.  The  information  should  be available
on  file  and provided  to USEPA within  7  days after  the  receipt of a written
request.

      Preliminary Results Delivery Schedule:

      If the sample arrives before 5  p.m.,  the  Preliminary Results  for that
      sample are  due within the  required turnaround time.   If the  sample is
      received after  5 p.m.,  the Preliminary Results for that sample are due
      within the  required turnaround  time beginning at 8 a.m. the  following
      day.  DELIVERABLES ARE  TO  BE REPORTED TOTAL AND  COMPLETE.   Concurrent
                                            B-7                                    OLC03.2

-------
Exhibit B -- Section 1
Contract Reports/Deliverables Distribution (Con't)

Footnotes (Con't):

      delivery is required.  Delivery shall be made such that all designated
      recipients receive the item on the same calendar day.  If the
      deliverables are due on a Saturday,  Sunday,  or Federal holiday,  then they
      shall  be delivered on the next business day.  Deliverables          delivered
      after  this time will be considered late.

      NOTE:  As specified in the Contract Schedule  (Government Furnished
      Supplies and Materials),  unless otherwise instructed by the CLP SMO based
      on a Regional decision, the Contractor  shall dispose of unused sample
      volume and used sample bottles/containers no earlier than sixty days
      following submission of the reconciled  Complete SDG File.  Sample
      disposal and disposal of unused sample  bottles/containers is the
      responsibility of the Contractor,  and should be done in accordance with
      all applicable laws and regulations governing disposal of such materials.
1.2    Distribution

       The following addresses correspond to the "Distribution" column in the
       tables in Sections 1.1.

        SMO:         USEPA Contract  Laboratory Program
                    Sample Management  Office  (SMO}1
                    2000 Edmund Halley Dr.
                    Reston,  VA 20191-3436

        Region:      USEPA Region: The  SMO will provide the Contractor with the
                    list of addresses  for the 10 USEPA Regions.   SMO will  provide
                    the Contractor  with updated Regional address/name lists as
                    necessary throughout the  period of the contract and identify
                    other client recipients on a case-by-case basis.

        USEPA AOC Organic Program Manager:

        Mailing Address:      USEPA OERR Analytical  Operations/
                              Data  Quality Center
                              Ariel Rios Building  (5204G)
                              1200  Pennsylvania Avenue,  N.W.
                              Washington, DC  20469
                              Attn: CLP Organic Program Manager

        Fed-Ex/Overnight      USEPA OERR Analytical  Operations/
        Delivery:             Data  Quality Center
                              1235  Jefferson  Davis Highway
                              Crystal  Gateway I,  12th Floor
                              Arlington, VA 22202
                              Attn: CLP Organic Program Manager
     JThe Sample Management Office (SMO) is a contractor-operated facility
operating under the  CLASS  contract awarded and administered by USEPA.

OLC03.2                                    B-8

-------
                                                                  Exhibit B -- Section  1
                                      Contract Reports/Deliverables Distribution  (Con't)

       USEPA Regional CLP Project Officer (CLP PO):

                SMO will provide the Contractor with the list of addresses for
                the CLP POs.  SMO will provide the Contractor with updated
                name/address lists as necessary throughout the period of the
                contract.

       QATS        USEPA Contract Laboratory Program  (CLP)
                   Quality Assurance  Technical Support  (QATS)  Laboratory2
                   2700 Chandler Avenue,  Building  C
                   Las Vegas, NV  89120
                   Attn:  Data Audit Staff
     2The  Quality Assurance Technical  Support  (QATS)  Laboratory is  a
contractor operated facility operating under the QATS contract awarded and
administered by the USEPA.

                                           B-9                                    OLC03.2

-------
Exhibit B -- Section 2
Reporting Requirements/Order of Data Deliverables

2.0    REPORTING REQUIREMENTS AND ORDER OF DATA DELIVERABLES

2.1    Introduction

       The Contractor shall provide reports and other deliverables as specified
       in the Contract Schedule (Performance/Delivery Schedule).   The required
       content and form of each deliverable is described in this exhibit.  All
       reports and documentation must be:

       •    Legible;

       •    Clearly  labeled and completed  in accordance  with instructions in
            this  exhibit;

       •    Arranged in the order  specified in this  section;

       •    Paginated consecutively in ascending order starting from the Sample
            Delivery Group (SDG) Narrative;

       •    Copies must be legible and double-sided;  and

       •    Information reported on the forms listed in  Exhibit B (excluding the
            Sample Log-in Sheet [DC-1]  and the Complete  SDG File  (CSF)  Inventory
            Sheet [DC-2])  must be  either typewritten or  computer-generated.
            Handwritten corrections of the information must be legible,  signed,
            and dated.

       NOTE: Complete SDG files need not be double-sided.   (The CSF is composed
       of original documents.)  However, Sample Data Packages delivered to
       Sample Management Office (SMO) must be double-sided.

2.1.1     Requirements for each deliverable item cited  in  the Contract  Schedule
           (Contract Performance/Delivery  Schedule)  are  specified in Sections
          2.3  to 2.9.   Prior to submission,  the Contractor shall.arrange items
          and  the components of each item in the order  listed in these
          sections.

2.1.2     The  Contractor shall use USEPA  Case numbers  (including SDG numbers)
          and  EPA sample numbers  to identify samples received under this
          contract,  both verbally and in  reports/correspondence.   The  contract
          number shall be specified in all correspondence.

2.2    Resubmission of Data

       If submitted documentation does not conform to the above criteria, the
       Contractor shall resubmit such documentation with deficiency(ies)
       corrected, at no additional cost to USEPA.

2.2.1     The  Contractor shall respond within seven days to written requests
          from data recipients for additional information  or explanations that
          result from the Government's inspection activities,  unless otherwise
          specified in the contract.

2.2.2     Whenever  the Contractor is required to submit or resubmit data as a
          result of an on-site laboratory evaluation, or through a Contract
          Laboratory Program Project Office (CLP PO)  action,  or  through a
          Regional  data reviewer's request,  the data shall be clearly  marked as
          ADDITIONAL DATA and shall be sent to both contractual  data recipients
           (SMO and  the Region; to USEPA designated  recipient,  upon written  •
          request).   The Contractor shall include a cover  letter which
          describes which data are being  delivered,  to  which USEPA Case(s)  the
          data pertain,  and who requested the data.


OLC03.2                                    B-10

-------
                                                                   Exhibit B -- Section 2
                                Reporting Requirements/Order of Data Deliverables  (Con't)

2.2.3     Whenever the Contractor is required to submit or resubmit data as a
          result of Contract Compliance Screening  (CCS) review by SMO, the data
          shall be sent to both contractual data recipients  (SMO and the
          Region; to the USEPA designated recipient  (e.g., QATS), when a
          written request for the Sample Data Package has been made).  In all
          instances, the Contractor shall include a color-coded COVER SHEET
           (Laboratory Response to Results of Contract Compliance Screening)
          provided by SMO.

2.3    Quality Assurance Plan (QAP)  and Standard Operating Procedures  (SOPs)

       The Contractor shall  adhere to the requirements in Exhibits E and F.

2.4    Sample Traffic Reports (TRs)

       Each sample received by the Contractor will be labeled with a EPA sample
       number, and will be accompanied by a Sample TR bearing the sample number
       and descriptive information regarding the sample.  The Contractor shall
       complete the TR (marked "Lab Copy for Return to SMO"), recording the
       date of sample receipt and sample condition upon receipt for each
       container, and shall sign the TR.  Information shall be recorded for
       each sample in the SDG.

2.4.1     The Contractor  shall submit  TRs in^SDG sets  (i.e., TRs  for all
          samples in an SDG  shall be clipped together), with an SDG cover sheet
          attached.  The  SDG cover  sheet shall  contain  the following items:

           •    Laboratory name;

           •    Contract number;

           •    Sample analysis price (full sample price from the contract);

           •    Case number; and

           •    List of EPA sample numbers of all samples in the SDG, identifying
               the first and last samples received, and their dates of receipt
               (LRDs).

          NOTE:  When more than one  sample  is received in  the first  or  last SDG
           shipment,  the  "first"  sample received would be  the lowest sample
          number (considering both  alpha and numeric designations); the  "last"
           sample received would  be  the highest  sample number (considering both
           alpha  and numeric  designations).

2.4.2      Each TR shall be  clearly  marked  with  the SDG number,  entered below
           the laboratory  receipt date  on the TR.   The TR  for the  last  sample
           received  in the SDG shall be clearly  marked "SDG--FINAL SAMPLE".   The
           SDG number is  the  EPA  sample number of  the first  sample received in
           the SDG.   When  several samples are received together in the  first  SDG
           shipment,  the  SDG  number  shall be the lowest sample  number
           (considering both  alpha and  numeric designations)  in the  first group
           of samples received under the SDG.

2.4.3      If samples are  received at the laboratory with  multi-sample  TRs, all
           the samples on one multi-sample  TR may not necessarily be in the same
           SDG.  In this instance, the  Contractor shall make  the appropriate
           number of photocopies  of  the TR,  and  submit one copy with each SDG
           cover  sheet.
                                           B-ll                                    OLC03.2

-------
Exhibit B -- Section 2
Reporting Requirements/Order of Data Deliverables  (Con't)

2 .5    Sample Data Package

       The Sample Data Package is divided into the five major units described
       in this section.  The last three units are each specific to an
       analytical fraction (volatiles, semivolatiles, and pesticides/Aroclors).
       If the analysis of a fraction is not required, then that
       fraction-specific unit is not required as a deliverable.  The Sample
       Data Package shall include data for the analyses of all samples in one
       SDG,  including field samples, dilutions, re-analyses, blanks. Laboratory
       Control Samples, and any requested Matrix Spike/Matrix Spike
       Duplicate(s) (MS/MSD).  The Contractor shall retain a copy of the CSF
       for 365 days after final acceptance of data.  After this time, the
       Contractor may dispose of the package.

2.5.1     SDG Narrative.  This document  shall  be  clearly labeled  "SDG
          Narrative" and  shall contain:  laboratory name;  Case  number,- EPA
          sample  numbers  in  the  SDG,  differentiating between initial analyses
          and re-analyses,- SDG number; Contract number,-  and detailed
          documentation of any quality control, sample,  shipment  and/or
          analytical problems encountered  in processing  the samples reported  in
          the data package.  The Contractor  shall include any  technical  and
          administrative  problems encountered, the corrective  actions taken,
          the resolution, and an explanation for  all flagged edits  (e.g.,
          manual  edits) on quantitation  lists.  This includes  documenting  the
          alternative  technique  used  to  determine cooler temperature if  a
          temperature  indicator  bottle is  not  present  in the cooler.  The
          Contractor shall also  provide, in  the SDG  Narrative,  sufficient
          information, including equations or  curves (at least one  equation or
          curve per method), to  allow the  recalculation  of sample results  from
          raw instrument  output.  The Contractor  shall also include a
          discussion of any  flexibility  SOW  modification.   This includes
          attaching a  copy of the approved modification  form to the SDG
          Narrative.   Additionally  the Contractor shall  also identify and
          explain any  differences which  exist  between  the Form Is and
          supporting documentation  provided  in the data  package and those
          previously provided as preliminary results.

          All Gas Chromatograph  (GC)  columns used for  analysis shall be
          documented here, by fraction.  List  the GC column identification--
          brand name,  the internal  diameter, in millimeters (mm), and the
          length,  in meters, packing/coating material  and film thickness.  The
          trap  used for volatile analysis  shall be described here.  List trap
          name, when denoted by  manufacturer,  its composition  [packing
          material/brand  name, amount of packing  material,  in  length,
          centimeters  (cm)].  All tentatively  identified (semi-volatile)
          alkanes and  their  estimated concentrations are to be reported  here.
          The EPA sample  number, the  CAS number,  when  available,  the alkane
          compound (or series) name,  and its estimated concentration shall be
          provided in  tabular format.  The Contractor  shall document in  the SDG
          Narrative all instances of  manual  integration.   The  SDG Narrative
          shall contain the  following statement,  verbatim:  "I  certify that this
          data  package is in compliance  with the  terms and conditions of the
          contract, both  technically  and for completeness,  for other than  the
          conditions detailed above.  Release  of  the data contained  in this
          hardcopy data package  and in the computer-readable data submitted on
          diskette has been  authorized by  the  laboratory manager  or his/her
          designee, as verified  by  the following  signature."   This  statement  .
          shall be directly  followed  by  an original  signature  of  the laboratory
          manager or their designee with a typed  line  below it containing  the
          signer's name and  title,  and the date of signature.
OLC03.2                                    B-12

-------
                                                                   Exhibit B -- Section 2
                                Reporting Requirements/Order of Data Deliverables  (Con't)

2.5.1.1       The  samples analyzed under  this  contract  should not  exhibit  a
              matrix effect  which would prevent  the  Contractor from meeting the
              requirements of  the contract.  Sample  re-extraction/re-analyses
              performed as a result of  suspected matrix interferences  beyond the
              scope of the method will  be reviewed on a case-by-case basis for
              payment purposes by the CLP PO.  Send  or  telefax to  the  CLP  PO a
              copy of the SDG  Narrative (include your contract number),  a
              description of the situation and the requested CLP PO action,
              either prior or  concurrent  with  the delivery of the  Sample Data
              Package.

2.5.1.2       The  Contractor shall list the pH determined for each water sample
              submitted for  volatiles analysis.   This information  may  appear as
              a simple list  or table in the SDG  Narrative.   The purpose of this
              pH determination is to ensure that all water volatiles samples
              were acidified in the field.   No pH adjustment is to be  performed
            .  by the Contractor on water  samples for volatiles analysis.

2.5.2     Traffic  Reports.  The Contractor  shall  include a  copy of  the  TRs
          submitted in Section 2.4  for all  of the samples in:the  SDG.   The  TRs
          shall be arranged in increasing EPA sample number order,  considering
          both letters and  numbers.  Copies of  the  SDG cover sheet  are  to be
          included with the copies  of  the TRs.   (See Section 2.4  for more
          detail on reporting requirements  for  TRs.)   In the case of multi-
          sample TRs,  the Contractor shall  make the appropriate number  of
          photocopies of the  TR so that  a copy  is submitted with  each
          applicable data package.  In addition,  in any instance  where  samples
          from more than one  multi-sample TR are in the same data package,  the
          Contractor shall  submit a copy of the SDG cover sheet with copies of
          the TRs.

2.5.3     Volatiles Data

2.5.3.1       Volatiles QC Summary

2.5.3.1.1        Deuterated Monitoring Compound Recovery  (Form II LCV-1,  LCV-2).

2.5.3.1.2        Matrix Spike/Matrix Spike  Duplicate Recovery (Form III LCV).
                 This data  shall be provided upon Region's  request.

2.5.3.1.3        Method Blank Summary  (Form IV  LCV): If more than a single  form
                  is necessary,  forms shall  be arranged in chronological order by
                 date of analysis of the blank,  by  instrument.

2.5.3.1.4        GC/MS instrument performance check (Form V LCV): If more than a
                  single form  is necessary,  forms shall be arranged  in
                  chronological order,  by instrument.

2.5.3.1.5         Internal Standard Area  and RT  Summary (Form VIII LCV): If  more
                  than a single form is necessary, forms shall be  arranged in
                  chronological order,  by instrument.

2.5.3.2       Volatiles  Sample  Data.   Sample data,  including dilutions, and re-
              analyses  data,  shall be  arranged  in packets with  the Organic
              Analysis  Data Sheet  (Form  I LCV-1, LCV-2, including Form  I  LCV-
              TIC),  followed  by the raw  data for volatile samples.  These sample
              packets  shall be  placed  in increasing EPA sample  number order,
              considering both  letters and numbers.

2.5.3.2.1         Target Compound Results,  Organic Analysis  Data Sheet  (Form I
                  LCV-1, LCV-2).  Tabulated results  (identification and
                  quantitation)  of the  specified target compounds  (Exhibit C,


                                           B-13                                    OLC03.2

-------
Exhibit B -- Section 2
Reporting Requirements/Order of Data Deliverables  (Con't)

                 Volatiles) shall be included.  The validation and  release of
                 these results are authorized by a specific,  signed statement  in
                 the SDG Narrative  (Section 2.5.1).  In  the event that  the
                 laboratory manager cannot verify  all data reported for each
                 sample, the laboratory manager shall provide a detailed
                 description of the problems associated  with  the sample in the
                 SDG Narrative.

2.5.3.2.2        Tentatively Identified Compounds  (Form  I LCV-TIC).   Form I
                 LCV-TIC.is the tabulated list of  the highest probable  match for
                 up to 30  organic compounds that are not deuterated monitoring
                 compounds or internal standard compounds and are not listed in
                 Exhibit C.  It includes the Chemical Abstracts Service (CAS)
                 registry  number  (if applicable),  tentative identification, and
                 estimated concentration.  This form shall be included  even if
                 no compounds are found.  If no compounds are found,  indicate
                 this on the form by entering  "0"  in the field for  "Number
                 Found".

2.5.3.2.3        .Reconstructed Total Ion Chromatograms  (for each sample or
                 sample extract, including dilutions and re-analyses).
                 Reconstructed ion chromatograms shall be normalized to the
                 largest nonsolvent component  and  shall  contain the following
                . header information:

                 •    EPA  sample number;

                 •    Date and time of analysis;
                                                       »
                 •    GC/MS instrument identifier;

                 •    Lab  file identifier,  and

                 •    Analyst ID.

2.5.3.2.3.1          Internal standards and deuterated monitoring compounds shall
                     be labeled with the names of  compounds,  either directly out
                     from the peak or on a printout of retention times,  if
                     retention times (or scan numbers)  are printed above the
                     peak.

2.5.3.2.3.2          If automated data system procedures are used for preliminary
                     identification and/or quantitation of the target compounds,
                     the complete data system report shall be included in all
                     Sample Data Packages,  in addition to the reconstructed ion
                     chromatogram.   The complete data system report shall include
                     all of the information listed below.  For laboratories which
                     do not use the automated data system procedures, a
                     laboratory "raw data sheet" containing the following
                     information shall be included in the'Sample Data Package,  in
                     addition to the chromatogram:

                     •  EPA sample number;

                     •  Date and time of analysis;

                     •  Retention time or scan number of identified target
                        compounds;

                     •  Ion used for quantitation  with measured area;

                     •  Copy of area table from data system;


OLC03.2                                    B-14

-------
                                                                   Exhibit B -- Section 2
                                Reporting Requirements/Order of Data Deliverables  (Con't)

                     •   GC/MS instrument identifier;

                     •   Lab file identifier;  and

                     •   Analyst ID.

2.5.3.2.3.3          In  all  instances where the  data system  report  has been
                    • edited,  or where manual  integration or  quantitation has been
                     performed,  the GC/MS  operator  shall identify such edits or
                     manual  procedures by  initialing and dating the changes made
                     to  the  report, and  shall include the  integration scan range.
                     In  addition,  a hardcopy  printout of the Extracted Ion
                     Current Profile  (EICP) of the  quantitation ion displaying
                     the manual integration shall be included in the raw data.
                     This  applies  to all compounds  listed  in Exhibit C
                     (Volatiles),  internal standards and deuterated monitoring
                     compounds.

                     •   EICPs displaying each manual integration.


2.5.3.2.4        Other Required Information.  For each sample, by each  compound
                 identified, the  following items shall be included  in the  data
                 package.


                 •    Copies of raw spectra  and copies of  background-subtracted
                      mass spectra of  target  compounds listed in Exhibit C
                      (Volatiles) 4,that are identified in the sample and
                      corresponding background-subtracted target compound
                      standard mass spectra.   Spectra shall  be labeled with EPA
                      sample number,  lab file identifier,  date and time of
                      analysis, and GC/MS instrument identifier.  Compound names
                      shall be clearly marked on all spectra.

                 •    Copies of mass spectra of non-deuterated monitoring/non-
                      internal standard organic compounds not listed in Exhibit C
                      with associated best-match spectra (minimum of one, maximum
                      of three best matches).  Spectra shall be labeled with EPA
                      sample number,  lab file identifier,  date and time of
                      analysis, and GC/MS instrument identifier. Compound names
                      shall be clearly marked on all spectra.

2.5.3.3       Volatiles Standards Data

2.5.3.3.1        Initial  calibration data (Form VI  LCV-1, LCV-2, LCV-3)  shall  be
                 included in order by instrument,  if more than one  instrument  is
                 used.


                 •    Volatile standard(s) reconstructed ion chromatograms and
                      quantitation reports for the  initial  (five-point)
                      calibration, labeled as in Section 2.5.3.2.3.  Spectra are
                      not required.

                 •    All initial calibration data  that pertain to  samples in the
                      data package shall be  included, regardless of when it was
                      performed and for which Case.  When more than one initial
                      calibration is performed, the data shall be in
                      chronological order, by instrument.

                 •    EICPs displaying each  manual  integration.


                                           B-15                                   "OLC03.2

-------
Exhibit B -- Section 2
Reporting Requirements/Order of Data Deliverables  (Con't)

2.5.3.3.2        Continuing calibration data  (Form VII LCV-1, LCV-2, LCV-3)
                 shall be  included in order by instrument, if more than one
                 instrument is used.

                 •    Volatile standard(s) reconstructed ion chromatograms and
                      quantitation reports for all continuing (12-hour)
                      calibrations,  labeled as in Section 2.5.3.2.3.   Spectra are
                      not required.

                 •   'When more than one continuing calibration is performed,
                      forms shall be in chronological order,  by instrument.

                 •    EICPs displaying each manual integration.

2.5.3.3.3        In all  instances where the data  system  report has been edited,
                 or where  manual integration or quantitation has been  performed,
                 the GC/MS operator shall  identify such  edits or manual
                 procedures by initialing  and dating the changes made  to  the
                 report, and  shall include the integration scan range.  In
                 addition, a  hardcopy printout of  the EICP of the quantitation
                 ion displaying the manual integration shall be included  in  the
                 raw data.  This applies  to all compounds listed in Exhibit  C
                  (Volatiles), internal standards  and deuterated monitoring
                 compounds.

2.5.3.4       Volatiles Raw QC Data

2.5.3.4.1        4-Bromofluorobenzene  (BFB) data  shall be arranged in
                 chronological order by instrument for each 12-hour period,  for
                 each GC/MS system utilized.

                 •    Bar graph spectrum,   labeled as in Section 2.5.3.2.4.

                 •    Mass listing,  labeled as in Section 2.5.3.2.4.

                 •    Reconstructed total  ion chromatogram,  labeled as in Section
                      2.5.3.2.3.

2.5.3.4.2        Blank data shall be arranged by  type of blank  (method, storage,
                 instrument)  and shall be in chronological order by instrument.

                 NOTE: This order is different from that used for samples.

                 •    Tabulated results (Form I LCV-1,  LCV-2).

                 •    Tentatively identified compounds (Form I LCV-TIC) even if
                      none are found.

                 •    Reconstructed ion chromatogram(s)  and quantitation
                      report(s),  labeled as in Section 2.5.3.2.3.

                 •    Target compound spectra with laboratory-generated standard
                      spectra, labeled as  in Section 2.5.3.2.4.   Data  systems
                      which are incapable  of dual display shall provide spectra
                      in the following order:

                      Raw  target compound spectra.

                      Enhanced or background-subtracted  spectra.

                      Laboratory-generated standard spectra.
OLC03.2                           ,         B-16

-------
                                                                   Exhibit B  --  Section 2
                                Reporting Requirements/Order of Data Deliverables (Con't)
                      GC/MS library search spectra for tentatively identified
                      compounds, labeled as in Section 2.5.3.2.4.

                      Quantitation/calculation of tentatively identified compound
                      concentrations.
2.5.3.4.3        Volatiles  Matrix Spike Data

                 •    Tabulated results (Form I LCV-l, LCV-2) of target
                      compounds.  Form I LCV-TIC is not required.

                 •    Reconstructed ion chromatogram(s) and quantitation
                      report(s), labeled as in Section 2.5.3.2.3.  Spectra are
                      not required.

2.5.3.4.4'       Volatiles  Matrix Spike Duplicate  Data

                 •    Tabulated results (Form I LCV-l., LCV-2) of target
                      compounds.  Form I LCV-TIC is not required.

                 •    Reconstructed ion chromatogram(s) and quantitation
                      report(s), labeled as in Section 2.5.3.2.3.  Spectra are
                      not required.

2.5.4      Semivolatiles Data
                                                              /•
2.5.4.1       Semivolatiles QC Summary

2.5.4.1.1        Deuterated Monitoring Compound Recovery (Form II LCSV-1,  LCSV-2) .

.2.5.4.1.2        Matrix Spike/Matrix Spike Duplicate Recovery (Form III  LCSV):
                 This data  shall be provided upon Region's request.

2.5.4.1.3        Method Blank Summary (Form IV LCSV):  If more than a single form
                  is necessary,  forms shall be arranged in chronological  order by
                 date of analysis of the  blank,  by instrument.

2.5.4.1.4        GC/MS Instrument Performance Check (Form V LCSV): If more than
                  a single form is necessary,  forms shall be arranged in
                  chronological order, by  instrument.

2.5.4.1.5         Internal Standard Area and RT Summary (Form VIII LCSV):  If more
                  than a single form is necessary,  forms shall be arranged in
                  chronological order, by  instrument.

2.5.4.2       Semivolatiles Sample Data.   Sample data,  including dilutions and
              re-analysis samples,  shall be arranged in packets  with the  Organic
              Analysis  Data Sheet  (Form I  LCSV-1,  LCSV-2,  including  Form  I
              LCSV-TIC),  followed by the raw  data  for  semivolatile samples.
              These sample  packets  shall be placed in  increasing EPA sample
              number order,  considering both  letters and  numbers.

 2.5.4.2.1         Target Compound Results,  Organic Analysis Data Sheet (Form  I
                  LCSV-1, LCSV-2).  Tabulated results  (identification and
                  quantitation) of the specified target compounds  (Exhibit C,
                  Semivolatiles) shall be  included.  The validation and release
                  of these results are authorized by a specific,  signed statement
                  in the SDG Narrative (Section 2.5.1).  In the event that the
                  laboratory manager cannot verify all data reported for each
                  sample, the laboratory manager shall provide a detailed
                                            B-17                                    OLC03.2

-------
Exhibit B -- Section 2
Reporting Requirements/Order of Data Deliverables  (Con't)

                 description of the problems associated with the sample in the
                 SDG Narrative.

2.5.4.2.2       _Tentatively Identified Compounds  (Form I LCSV-TIC).  Form I
                 LCSV-TIC is the tabulated  list of the highest probable match
                 for up to 30 organic compounds that are not deuterated
                 monitoring compounds, or internal standard organic compounds
                 and are not listed in Exhibit C  (Volatiles, Semivolatiles).  It
                 includes the CAS registry  number  (if applicable), tentative
                 identification, and estimated concentration.  This form shall
                 be included even if no compounds  are found.  If no compounds
                 are found, indicate this on the form by entering  "0" in the
                 field for "Number TICs found."

2.5.4.2.3        Reconstructed Total Ion Chromatograms  (for each sample,
                 including dilutions and re-analyses).  Reconstructed ion
                 Chromatograms shall be normalized to the largest  nonsolvent
                 component and shall contain the following header  information:

                 •    EPA sample number;

                 •    Date and time of analysis;

                 •    GC/MS instrument identifier;

                 •    Lab file identifier;  and

                 •    Analyst ID.

2.5.4.2.3.1          Internal standards and deuterated monitoring compounds shall
                     be labeled with the  names of compounds,  either directly  out
                     from the peak or on a printout of retention times  if
                     retention times (or  scan numbers)  are printed over the peak.

2.5.4.2.3.2          If automated data system procedures are used for preliminary
                     identification and/or quantitation of the  target compounds,
                     the complete  data system report shall' be included  in  all
                     Sample Data Packages,  in addition to the reconstructed ion'
                     chromatogram.   The complete data system report shall  include
                     all of the information listed below.   For laboratories which
                     do not use the automated data system procedures, a
                     laboratory "raw data sheet"  containing the following
                     information shall be included in the Sample Data Package,  in
                     addition to the chromatogram:

                     •  EPA sample number;

                     •  Date and time of analysis,-

                     •  Retention time or scan number of identified target
                        compounds;

                     •  Ion used for quantitation with measured area;

                     •  Copy of area table  from data system;

                     •  GC/MS instrument identifier;

                     •  Lab file identifier; arid

                     •  Analyst ID.
 OLC03.2                                    B-18

-------
                                                                   Exhibit B -- Section 2
                                Reporting Requirements/Order of Data Deliverables (Con't)

2.5.4.2.3.3         In all instances where the data system report has been
                    edited, or where manual integration or quantitation has been
                    performed, the GC/MS operator shall identify such edits or
                    manual procedures by initialing and dating the  changes made
                    to the report, and shall include the integration scan range.
                    In addition, a hardcopy printout of the EICP of the
                    quantitation ion displaying the manual integration shall be
                    included  in the raw data.  This applies to all  compounds
                    listed in Exhibit C  (Semivolatiles), internal standards and
                    deuterated monitoring compounds.

                    •   EICPs displaying each manual integration.

2.5.4.2.4        Other Required Information.  For each sample, by each compound
                 identified,  the following shall be included in the data
                 package.

                 •    Copies  of  raw spectra  and copies  of background-subtracted
                      mass spectra  of  target compounds  listed in Exhibit  C
                      (Semivolatiles)  that are  identified in the sample and
                      corresponding background-subtracted target compound
                      standard mass spectra.   Spectra  shall  be labeled with EPA-
                      sample  number,  lab file identifier, and date  and time of
                      analysis,  and GC/MS instrument identifier compound  names
                      shall be clearly marked on all spectra.

                 •    Copies  of  mass spectra of non-deuterated monitoring
                      compounds/non-internal standard  organic compounds not
                      listed  in  Exhibit  C (Volatiles and Semivolatiles) with
                      associated best-match  spectra (minimum of- one, maximum of
                      three best matches).   This includes the mass  spectra for
                      tentatively identified alkanes.   Spectra shall be  labeled
                      with EPA sample-number,  lab file identifier,  and  date and
                      time of analysis,  and  GC/MS instrument identifier  compound
                      names shall be clearly marked on all  spectra.

2.5.4.3       Semivolatiles Standards Data

2.5.4.3.1        Initial  calibration data (Form VI LCSV-1,  LCSV-2,  LCSV-3) shall
                 be included  in order by instrument, if more  than one instrument
                 is used.

                 •    Semivolatile standard(s), reconstructed ion chromatograms,
                      and quantitation reports for the initial (five-point)
                      calibration,  labeled as in Section 2.5.4.2.3.   Spectra are
                      not required.

                 •    All initial calibration data that pertain to  samples in  the
                      data package shall be included,  regardless of when  it was
                      performed and for which Case.  When more than one initial
                      calibration is performed, the data shall be in
                      chronological order, by instrument.

                 •    EICPs displaying each manual integration.

2.5.4.3.2        Continuing  calibration data (Form VII LCSV-1, LCSV-2, LCSV-3)
                 shall be included in order by instrument,  if more  than  one
                 instrument  used.

                  •    Semivolatile standard(s) reconstructed ion chromatograms
                      and quantitation reports for all continuing (12-hour)
                                           B-19                                    OLC03.2

-------
Exhibit B -- Section 2
Reporting Requirements/Order of Data Deliverables  (Con't)

                      calibrations,  labeled as in Section 2.5.4.2.3.  Spectra are
                      not required.

                 •    When more than one continuing calibration is performed,
                      forms shall be in chronological order,  by instrument.

                 •    EICPs displaying each manual integration.

2.5.4.3.3        In all  instances where the data  system  report has  been edited,
                 or where manual integration or guantitation has been performed,
                • the GC/MS operator shall identify such  edits or manual
                 procedures by initialing and dating the changes made to the
                 report, and shall include the integration scan range.  In
                 addition, a hardcopy printout of  the EICP of the quantitation
                 ion displaying the manual integration shall be included in  the
                 raw data.  This applies to all compounds listed in Exhibit  C
                  (Semivolatiles), internal standards, and deuterated monitoring
                 compounds.

2.5.4.4       Semivolatiles Raw QC Data

2.5.4.4.1        Decafluorotriphenylphosphine  (DFTPP) data shall be arranged in
                 chronological order by instrument for each 12-hour period,  for
                 each GC/MS system utilized.

                 •    Bar graph spectrum,  labeled as in Section 2.5.4.2.4.

                 •    Mass listing,  labeled as in Section 2.5.4.2.4.

       fO.        •    Reconstructed total ion chromatogram,  labeled as in Section
                      2.5.4.2.3.

2.5.4.4::'2        Blank data shall be included in  chronological order by
       ••;        extraction date.

                 NOTE: This order is different from that used  for samples.

                 •    Tabulated results (Form I LCSV-1,  LCSV-2).

                 •    Tentatively identified compounds (Form I LCSV-TIC)  even if
                      none are found.

                 •    Reconstructed ion chromatogram(s)  and quantitation
                      report(s),  labeled as in Section 2.5.4.2.3.

                 •    Target compound spectra with laboratory-generated standard
                      spectra,  labeled as in Section 2.5.4.2.4.  Data systems
                      which are incapable of dual display shall provide spectra
                      in the following order:

                      Raw target compound spectra.

                      Enhanced or background-subtracted  spectra.

                      Laboratory-generated standard spectra.

                 •    GC/MS library search spectra for tentatively identified
                      compounds,  labeled as in Section 2.5.4.2.4.

                 •    Quantitation/calculation of tentatively identified compound
                      concentrations.
OLC03.2                                    B-20

-------
                                                                   Exhibit B -- Section 2
                                Reporting Requirements/Order of Data Deliverables (Con't)

2.5.4.4.3        Semivolatiles Matrix Spike Data

                 •    Tabulated results  (Form I  LCSV-1,  LCSV-2)  of  target
                      compounds.   Form I  LCSV-TIC is  not required.

                 •    Reconstructed ion chromatogram(s)  and  quantitatioh
                      report(s),  labeled  as in Section  2.5.4.2.3.   Spectra are
                      not  required.
2.5.4.4.4        Semivolatiles Matrix Spike Duplicate Data

                 •    Tabulated results  (Form I  LCSV-l,  LCSV-2)  of  target
                      compounds.   Form I  LSV-TIC is not required.

                 •    Reconstructed ion chromatogram(s)  and  quantitation
                      report(s),  labeled  as in Section  2.5.4.2.3.   Spectra are
                      not  required.

2.5.5      Pesticide/Aroclor Data

2.5.5.1       Pesticide/Aroclor QC Summary

2.5.5.1.1        Surrogate  Percent Recovery  Summary  (Form II LCP).

2.5.5.1.2        Matrix Spike/Matrix  Spike Duplicate Recovery  (Form III LCP-l):
                 This data  shall be provided upon  Region's request.

2.5.5.1.3        Laboratory Control Sample Recovery  (Form III LCP-2).

2.5.5.1.4        Method Blank  Summary (Form  IV  LCP): If more than  a single  form
                 is necessary, forms  shall be arranged  in chronological order  by
                 date of  analysis of  the  blank.

2.5.5.2       Pesticide/Aroclor Sample Data.  Sample  data including dilutions
              and re-analyses shall be arranged in packets with the Organic
              Analysis Data Sheet (Form I LCP) and the raw data for pesticide
              samples.  These sample packets should then be placed in increasing
              .EPA sample number order, considering both letters and numbers.

2.5.5.2.1        Target Compound  Results,  Organic  Analysis Data Sheet  (Form I
                 LCP).  Tabulated results (identification and quantitation)  of
                 the  specified target compounds (Exhibit C,  Pesticides/Aroclors)
                 shall be included.   The  validation  and release of these  results
                 is authorized by a  specific, signed statement  in  the  SDG
                 Narrative (see  Section  2.5.1).   In  the event that the
                 laboratory manager  cannot verify all  data  reported for each
                 sample,  the  laboratory  manager shall  provide a detailed
                 description  of  the  problems associated with the sample in the
                 SDG  Narrative.

2.5.5.2.2        Copies of Pesticide Chromatograms.  Positively identified
                  (identified  according to the criteria in Exhibit  D Pesticides
                 and  Aroclors)  compounds for each column shall  be  labeled with
                  the  names of compounds,  either directly out from  the  peak on
                  the  chromatogram,  or on a printout  of retention times on the
                  data system  printout if retention times are printed over the
                 peak on  the  chromatogram.   All chromatograms  shall meet  the
                  acceptance criteria in  Exhibit D Pesticides and Aroclors,  and
                  shall be labeled with the following information:

                  •   EPA sample number,-

                  •   Volume  injected (uL) ,-


                                           B-21                                    OLC03.2

-------
Exhibit B -- Section 2
Reporting Requirements/Order of Data Deliverables  (Con't)


                 •    Date and time of injection;

                 •    GC column identifier (by stationary phase and internal
                      diameter);

                 •    GC instrument identifier; and

                 •    Scaling factor.

2.5.5.2.3        Copies of pesticide chromatograms from  the second GC column
                 shall be included and labeled as  in Section 2.5.5.2.2.

2.5.5.2.4        Data System Printout.  A printout of retention time and
                 corresponding peak height or  peak area  shall accompany each
                 chromatogram.  The printout shall be labeled with the EPA
                 sample number.  In all instances where  the data system report
                 has been edited, or where manual  integration or quantitation
                 has been performed, the GC/ECD  operator must identify such
                 edits or manual procedures by initialing and dating the changes
                 made to the report, and shall include the integration time
                 range.

2.5.5.2.5        All manual work sheets shall  be included in the Sample Data
                 Package.

2.5.5.3       Pesticide/Aroclor Standards Data

2.5.5^.3.1        Initial Calibration of Single Component Analytes  (Form VI  LCP-
    •:'r           1, LCP-2) : for all GC columns,  all instruments, in
                 chronological order by GC column  and instrument.

2.5.5';;3.2        Initial Calibration of Multicomponent Analytes  (Form VI LCP-3) :
    -~           for all GC columns, all instruments, in chronological order by
     :::           GC column and instrument.

2.5.5-.3.3        Analyte Resolution Summary  (Form VI LCP-4): for all GC columns
                 and instruments, in chronological order by GC column and
                 instrument.

2.5.5.3.4        Performance Evaluation Mixture  (Form VI LCP-5): for all GC
                 columns and instruments, in chronological order by GC column
                 and instrument.

2.5.5.3.5        Individual Standard Mixture A (Form VI  LCP-6)-. for all GC
                 columns and instruments, in chronological order by GC column
                 and instrument.

2.5.5.3.6        Individual Standard Mixture B (Form VI  LCP-7): for all GC
                 columns and instruments, in chronological order by GC column
                 and instrument.

2.5.5.3.7        Calibration Verification Summary  (Form  VII LCP-1): for all
                 performance evaluation mixtures and instrument blanks, on  all
                 GC columns and instruments, in  chronological order by GC column
                 and instrument.  Report for each  associated instrument blank.

2.5.5.3.8        Calibration Verification Summary  (Form  VII LCP-2): for all mid-
                 point concentrations of Individual Standard Mixtures A and B
                 and reported for all instrument blanks  used for calibration
                 verification, on all GC columns and instruments, in
                 chronological order by GC column  and instrument.


OLC03.2                                    B-22

-------
                                                                   Exhibit B -- Section 2
                                Reporting Requirements/Order of  Data Deliverables (Con't)

2.5.5.3.9        Analytical Sequence (Form VIII LCP):  for all GC columns and
                 instruments,  in chronological order by GC column and
                 instrument.

2.5.5.3.10       Florisil Cartridge Check (Form IX LCP): for all lots of
                 cartridges used to process samples in the SDG.

2.5.5.3.11       Pesticide Identification Summary for Single Component Analytes
                 (Form X LCP-1): for all samples with positively identified
                 single component analytes,  in order by increasing EPA sample
                 number.

2.5.5.3.12       Pesticide Identification Summary for Multicomponent Analytes
                 (Form X LCP-2): for all samples with positively identified
                . multicomponent analytes, in order by increasing EPA sample
                 number.

2.5.5.3.13       Chromatograms and data system printouts shall be included for
                 all standards including the following:

                 •    Resolution check  mixture;

                 •    Performance evaluation mixtures,  all;

                 •    Individual Standard Mixture A,  at three concentrations,
                      each initial calibration;

                 •    Individual Standard Mixture B,  at three concentrations,
                      each initial calibration;                                           ,

                 •    All multicomponent analytes (toxaphene and Aroclors),  each
                      initial  calibration;

                 •    All mid-point concentrations of Individual Standard
                      Mixtures A and B  used for calibration verification; and

                 •    All multicomponent analyte standards analyzed for
                      confirmation.

2.5.5.3.14       A printout of retention  time and corresponding peak height or
                 peak area shall  accompany each  chromatogram.  The printout
                • shall be labeled with the EPA sample number.  In addition, all
                 chromatograms shall meet the acceptance criteria in Exhibit D
                 Pesticides and Aroclors, and shall be  labeled with the
                 following:

                 •    EPA sample number for the standard,  e.g.,  INDA,  INDB,  etc.
                      (See Section 3.3.7.6 for details);

                 •    Labels of all standard peaks for all individual compounds
                      either directly out from the peak on the chromatogram or on
                      the printout of retention times on the data system printout
                      if retention times are printed over the peak on the
                      chromatogram;

                 •    Total nanograms injected for each standard.  When total
                      nanograms injected appear on the printout, it is not
                      necessary to include them on the chromatogram;

                 •    Date and time of  injection;
                                           B-23                                    OLC03.2

-------
Exhibit B -- Section 2
Reporting Requirements/Order of Data Deliverables  (Con't)

                 •    GC column identifier (by stationary phase and internal
                      diameter);

                 •    GC instrument identifier; and

                 •    Scaling factor.

                 NOTE:.In all instances where  the  data system report has been
                 edited, or where manual  integration or quantitation has been
                 performed, the GC/EC operator must identify such edits or
                 manual procedures by initialing and dating the changes made to
                 the report, and shall include the integration time range.

2.5.5.4       Pesticide/Aroclor Raw QC Data

2.5.5.4.1        Blank data shall be arranged  by type of blank  (method,
                 instrument, sulfur cleanup) and shall be  in chronological  order
                 by instrument.

                 NOTE: This order is different from that used for samples.

                 •    Tabulated results (Form  I LCP).

                 •    Chromatogram(s)  and data system printout(s)  (GC)  for each
                      GC column and instrument used for analysis,  labeled as in
                      Sections 2.5.5.2.2 and 2.5.5.2.4.

2.5.5.4.2        Laboratory Control Sample Data

                 •    Tabulated results (Form  I LCP)  of target compounds.

                 •    Chromatogram(s)  and data system printout(s)  (GC),  labeled
                      as in Sections 2.5.5.2.2 and 2.5.5.2.4 and for both columns
    ;-"                as in Section 2.5.5.2.3.

2.5.5.4.3        Pesticides Matrix Spike                '

                 •    Tabulate results (Form I LCP) of target compounds.

                 •    Chromatogram(s)  and data system printout(s)  (GC),  labeled
                      as in Section 2.5.5.2.2  and 2.5.5.2.4 and for both columns
                      as in Section 2.5.5.2.3.

2.5.5.4.4        Pesticides Matrix Spike  Duplicate Data

                 •    Tabulate results (Form I LCP) of target compounds.

                 •    Chromatogram(s)  and data system printout(s)  (GC),  labeled
                      as in Section 2.5.5.2.2  and 2.5.5.2.4 and for both columns
                      as in Section 2.5.5.2.3.

2.5.5.5       Raw Florisil Data.  The chromatogram and data system report(s)
              shall be arranged in chronological order by Florisil cartridge
              performance check analyses.

              •    Chromatograms and data  system reports  labeled as  specified in
                   Sections  2.5.5.2.2  and  2.5.5.2.4 for the florisil  cartridge
                   performance  check analyses.

              •    Chromatograms and data  system reports  for standard analyses
                   used to quantify compounds in the Florisil  cartridge
                   performance  check analysis,  labeled as specified  in Section


OLC03.2                                    B-24

-------
                                                                   Exhibit B -- Section 2
                                Reporting Requirements/Order of Data Deliverables (Con't)

                  2.5.5.3.14 (i.e., Individual Standard Mixture A and Individual
                  Standard Mixture B and the 2,4,5 Trichlorophenol solution).

2.6    Complete SDG File (CSF)

       As specified in Section  1,  the  Contractor shall  deliver one Complete SDG
       File (CSF)  including the original Sample Data Package to the Region
       concurrently with delivery of the Sample Data Package to SMO.   (Delivery
       to a USEPA designated recipient is only required upon written request.)

2.6.1     The CSF will contain all original documentssas specified in Section  3
          and Exhibit F, and in Form DC-2  (see Section 4).   No photocopies of
          original documents will be placed in the CSF unless the original data
          was initially written in a bound notebook, maintained by the
          Contractor, or the originals were previously submitted to USEPA with
          another Case/SDG in accordance with the requirements described in
          Exhibit F.  The contents of the CSF shall be numbered according to
          the specifications described  in Section 3.20.

2.6.2     The CSF will consist of the following original documents in addition
          to the documents in the Sample Data Package.

          NOTE: All SDG-related documentation may be used or admitted as
          evidence in subsequent legal proceedings.  Any other SDG-specific
          documents generated after the CSF is sent to USEPA, as well as copies
          that  are altered in any fashion, are also deliverables to USEPA.'
           (Deliver the original to the  Region and a copy to SMO.  Delivery to  a
          USEPA-designated recipient  is only upon written request.)

2.6.2.1       The original Sample Data Package.

2.6*2.2       A completed and signed document  inventory sheet (Form DC-2).

2.6.2.3       All original shipping documents  including,  but not limited to,  the
              following documents:

              •   USEPA Chain-of-Custody Record;

              •   Airbills  (if  an  airbill  is not received,  include  a hardcopy
                   receipt  requested from the shipping  company or  a  printout  of
                   the shipping company's electronic  tracking  information);

              •   USEPA TRs; and

              •    Sample tags  (if  present)  sealed in plastic  bags.

2.6.2.4       All original receiving documents including, but not limited to,
              the following documents:

              •    Form DC-1;

              •    Other receiving forms or copies of receiving logbooks;  and

              •    SDG cover sheet.

2.6.2.5       All original laboratory records, not already submitted in the
              Sample Data Package, of sample transfer,  preparation and analysis
              including, but not limited to, the following documents:

              •    Original preparation and analysis forms  or copies of
                   preparation  and analysis logbook  pages;
                                           B-25                                    OLC03.2

-------
Exhibit B .-- Section 2
Reporting Requirements/Order of Data Deliverables  (Con't)

              •    Internal  sample and sample extract transfer chain-of-custody
                   records;

              •    Screening, records; and

              •    All  instrument output, including strip charts from  screening
                   activities.

2.6.2.6       All other original SDG-specific documents in the possession of the
              Contractor including,  but  not limited to, the following documents:

              •    Telephone contact  logs;

              •    Copies of personal logbook pages;

              •    All  handwritten SDG-specific notes;  and

              •    Any  other SDG-specific documents not covered by  the above.

2.6.3      If  the Contractor does submit SDG-specific documents to  USEPA after
           submission of the CSF, the documents should  be identified with unique
           accountable  numbers,  a revised Form DC-2 should  be  submitted,  and the
           unique accountable numbers and the'locations of  the documents in  the
           CSF should be recorded in  the "Other Records"  section on the  revised
           Form DC-2.   Alternatively, the Contractor may number the newly
           submitted SDG-specific documents to USEPA as a new  CSF and  submit a
           new Form DC-2.  The  revised Form DC-2  or new Form DC-2 should be
           submitted to the  USEPA Regions only.

2.7    /Data in Computer-Readable Format

       5The Contractor  shall provide a computer-readable  copy of the data on
       data reporting  Forms I-X for all samples in the SDG.as specified in
       ^Exhibit H,  and  delivered as specified in the Contract Schedule (Contract
       -Performance/Delivery Schedule).   Computer-readable data deliverables
       shall be  submitted on IBM or IBM-compatible, 3.5-inch high-density 1.44
       tMB diskette (or via an alternate means of electronic transmission
       approved  in advance by the USEPA) .

2.7.1      When submitted, the  diskette(s)  shall  be packaged and shipped in  such
           a manner that the diskette(s) cannot be  bent or  folded,  and will  not
           be  exposed to extreme heat or cold or  any type of electromagnetic
           radiation.   The diskette(s) shall be included in the same shipment as
           the hardcopy data and shall,  at  a minimum, be enclosed in a diskette
           mailer.   The diskette(s) shall be labeled as specified in Exhibit H,
           Section 8.4.

2.7.2      The data shall be recorded in ASCII, text file format, and  shall
           adhere to the file,  record, and  field  specifications listed in
           Exhibit H.

2.8    Preliminary Results

       The Form  I  data results shall be submitted  for all samples  in one SDG of
       a Case.   This includes tabulated target compound  results (Form I) for
       the volatile, semivolatile,  and pesticide fractions, and tentatively
       identified  compounds (Form I TIC) for the volatile and semivolatile
       fractions.  The Contractor shall clearly identify the Preliminary
       Results by  labeling each Form I and Form I TIC as"Preliminary Results"
       under each  form title (e.g.,  under Volatile Organics Analysis Data
       Sheet, Volatile Organics Analysis Data Sheet Tentatively Identified
       Compounds).


OLC03.2  '                                  B-26

-------
                                                                   Exhibit B -- Section 2
                                Reporting Requirements/Order of Data Deliverables (Con't)

2.8.1     The Contractor shall also include a disclaimer at the bottom of all
          Form Is stating that the "Data results contained on this Form I are
          for scanning purposes only and may not have been validated fro CLP
          criteria."                                                             .     •

2.8.2     Sample Traffic Reports and SDG Cover Sheets shall be submitted with
          the Preliminary Results.

2.9    GC/MS and GC/ECD Tapes

       The Contractor shall adhere to the requirements in Section 13 of
       Exhibit E.

2.10   Extracts

       The Contractor shall preserve sample extracts at 4°C (±2°C)  in
       bottles/vials with PTFE-lined septa.  Extract bottles/vials shall be
       labeled with EPA sample number,  Case number,  and SDG number.  The
       Contractor shall maintain a logbook of stored extracts,  listing EPA
       sample numbers and associated Case and SDG numbers.  The Contractor
       shall retain extracts for 365 days following submission of the
       reconciled complete Sample Data Package.  During that time,  the
       Contractor shall submit extracts and associated logbook pages within
       seven days following receipt of a written request from the CLP PO.
                                           B-27                                    OLC03.2

-------
Exhibit B -- Section 3
Form Instructions

3 . 0    FORM INSTRUCTIONS

3 .1    Introduction

       This section includes specific instructions for completing the data
       reporting forms required under this contract.   Each of the forms is
       specific to a given fraction (volatile,  semivolatile,  or
       pesticide/Aroclor).   The Contractor shall submit only those forms
       pertaining to the fractions analyzed for a given sample(s).   For
       instance, if a sample is scheduled for volatiles analysis only, the
       Contractor shall provide only forms for the volatile fraction.

3.2    General Information

       The Contractor shall report values on the hardcopy forms according to
       the individual form instructions in this section.  For instance, all
       results for concentrations of target compounds shall be reported to two
       significant figures.  Values that exceed the maximum length allowed
       shall be reported to the maximum possible, maintaining the specified
       significance.

3.2.1     The  data  reporting  forms presented in  Section 4 have been  designed  in
          conjunction with the  computer-readable data  format  specified in
          Exhibit H.  The  specific length  of each variable for
          computer-readable data  transmission purposes is also given in Exhibit
          H.   Information  entered on  these forms shall not exceed  the size of
          the  field given  on  the  form,  including such  laboratory-generated
          items as  lab  name and lab sample identifier.

          NOTE:  The space  provided for entries on the  hardcopy forms (Section
          4)  is greater in some instances  than the length prescribed for  the
          variable  as written to  the  electronic  deliverable  (see Exhibit  H).
          Greater space is provided on the hardcopy  forms for visual clarity.

3.2.2     When submitting  data, the Contractor shall reproduce all characters
          that appear on the  data reporting forms in Section  4.  The format of
          the  forms submitted shall be identical to  that shown in  the contract.
          No  information may  be added,  deleted,  or moved from its  specified
          position  without prior  written approval from the Contract  Laboratory
          Program Project  Officer (CLP PO).  The names of the various fields
          and  compounds (e.g.,  "Lab Code," "Chloromethane")  shall  appear  as
          they do on the forms  in the contract.   For items appearing on the
          uncompleted forms (Section  4), the use of  uppercase and  lowercase
          letters is optional.

3.2.3     Alphabetical  entries  made on the forms by  the Contractor shall  be in
          ALL  UPPERCASE letters (e.g.,  "ABCDE",  not  "Abcde"  or  "abcde").   If  an
          entry does not fill the entire blank space provided on the form, null
          characters shall be used to remove the remaining underscores that
          comprise  the  blank  line.    However, the Contractor  shall not remove
          the  underscores  or  vertical bars that  delineate  "boxes"  on the  forms.
          The  only  exception  would be those underscores at the bottom of  a
           "box" that are intended as  a data entry line.   (For instance, on Form
          2LCV,  line 30, if data  is entered on line  30, it will  replace the
          underscores.)
OLC03.2                                    B-28

-------
                                                                   Exhibit B -- Section 3
                                                                Form Instructions (Con't)
3.3    Header Information
       Six pieces of information are common to the  header  section of each data
       reporting form:  lab name,  contract,  lab code,  Case  number.  Client number
       and Sample Delivery Group (SDG)  number.   Except  as  noted for Client
       number,  this information shall be entered on every  form and shall match
       on every form.

3.3.1     Lab Name.  The lab name shall be the name chosen by the Contractor to
          identify its laboratory.   It shall not exceed 25 characters.

3.3.2     Contract.  Contract refers to the number of the USEPA contract under
          which the analyses were performed.

3.3.3     Lab Code.  -The lab code is an alphabetical abbreviation of up to six
          letters, as assigned bv USEPA. to identify the laboratory and aid in
          data processing.  This lab code will be assigned by USEPA at the time
          a contract is awarded, and shall not be modified by the Contractor,
          except  at the direction of USEPA.  If a change of name or ownership
          occurs  at the laboratory,  the lab code will remain the same until the
          Contractor is directed by USEPA to use another lab code.

3.3.4     Case Number.  The Case number is the USEPA-assigned Case number
          associated with  the sample.  This number is reported on the Traffic
          Report  (TR).

3.3.5     Client  Number.   The Client number is a unique number identifying the
          client  and the project.   This number may be the USEPA-assigned number
          for analyses performed under Non-Routine Analytical Services  (NRAS).
          If samples are  to be  analyzed under NRAS only, and reported on  these
          forms,  then enter the NRAS number as  "Client No." and  leave the Case
          number  blank.   If samples are analyzed according to the Routine
          Analytical Services  (RAS)  protocol and have additional NRAS
          requirements, list both the  Case number and NRAS number on all  forms.
          If the  analyses  have  no NRAS requirements, leave the  "Client No."
          field blank.

          NOTE:  Some samples in an  SDG may have a Client Number  where as  other
          may not.

3.3.6     SDG Number.  The "SDG No." is the Sample Delivery Group  (SDG) number.
          The  SDG number  is the EPA sample number of the  first  sample received
           in the  SDG,  except when this would  cause.duplication.  When several
           samples are  received  together in the  first SDG  shipment,  the  SDG
          number  shall be the  lowest sample number  (considering  both alpha and
          numeric designations)  in  the first  group of  samples  received  under
           the  SDG.   If fractions of the same  field samples are  scheduled under
           different turnaround  times,  thus creating  separate  SDG's  containing
           the  same sample numbers,  a different  sample-number  shall  be utilized
           in the assignment of  the  SDG number for  each SDG.   If  a  situation
           arises  where there are an insufficient number of samples  for
           assignment  of SDG numbers the Contractor  shall  contact SMO  for the
           assignment  of an SDG number.

3.3.7      Sample Number.   This  number appears either in the upper  right-hand
           corner of the form,  or as the left  column  of a  table  summarizing data
           from a number of samples.  When  the EPA  sample  number is entered in
           the  triple-spaced box in  the upper  right-hand corner of  Form  I, Form
           IV,  or Form X,  it should  be entered on the middle line of the three
           lines that comprise  the box.
                                           B-29                                    OLC03.2

-------
Exhibit B -- Section 3
Form Instructions  (Con't)

3.3.7.1       The Contractor shall identify all samples, including dilutions and
              re-analyses,  Laboratory Control Samples,  requested Matrix
              Spike/Matrix Spike Duplicate(s) (MS/MSD)   (as described in Section
              3.3.7.4),  blanks,  and standards with an EPA sample number.  For
              field samples, the EPA sample number is the five digit unique
              identifying number given in the TR that accompanied that sample.
              In order to facilitate data assessment, the Contractor shall use
              the following sample suffixes:

              XXXXX        =  EPA sample number
              XXXXXMS      =.  Matrix spike sample
              XXXXXMSD     =  Matrix spike duplicate sample
              XXXXXRE      =  Re-extracted and re-analyzed sample
              XXXXXDL      =  Sample analyzed at a dilution
              XXXXXDL2     =  Sample analyzed at a secondary dilution
              XXXXXDL3     =  Sample analyzed at a third dilution	

              NOTE: The Region may approve up to one additional dilution be
              performed beyond the one dilution for volatiles and semivolatiles
              and two dilutions for pesticides specified in Exhibit D.  The
              approval of the additional dilution by the Region must be
              documented in the SDG Narrative and include the Telephone Record
              Conversation between SMO and the Contractor communicating USEPA's
              decision.

3.3..7.2       There may be instances when all samples analyzed must be listed 'on
  • •:..          the form,  regardless of whether or not they are part of the SDG
  „:':. .         being reported (e.g.. Form VIII LCP)..   In these instances, use
  •;-..          ZZZZZ as the EPA sample number for any sample analysis not
  :..x          associated with the SDG being reported.
  .3'
3.3;.,7.3       For blanks, the Contractor shall use the  following identification
              scheme for the EPA sample number:

              •    Volatile method blanks shall  be identified as VBLK##;

              •    Volatile instrument blanks shall  be  identified as VIBLK##;

              •    Volatile storage blanks shall be  identified as VHBLK##;

              •    Semivolatile  method blanks shall  be  identified as SBLK##;

              •    Pesticide/Aroclor method blanks and/or sulfur cleanup blanks
                   shall  be identified as PBLK##;  and

              •    Pesticide/Aroclor instrument  blanks  shall  be identified as
                   PIBLK##.

3.3.7.3.1         The  EPA sample number shall be unique for each blank within  an
                  SDG.  Within a fraction,  the Contractor shall achieve  this by
                  replacing the  two-character terminator (##)  of the identifier
                  with one  or two characters or  numbers,  or a combination of
                  both.   For example,  possible identifiers for volatile  blanks
                  would be  VBLK1,  VBLK2,  VBLKA1, VBLKB2,  VBLK10,  VBLKAB,  etc.

3.3.7.3.2         If the  method  blank is analyzed  on multiple instruments,  then
                  an additional  two-character suffix shall be added to make the
                  blank EPA sample number unique.
OLC03.2                                    B-30

-------
                                                                   Exhibit B -- Section 3
                                                                Form Instructions (Con't)
3.3.7.4      The  EPA sample  number  shall be unique  for  each Laboratory Control
             Sample  within an SDG.  The EPA sample  number  for  a Laboratory
             Control Sample  must  be PLCS##.

             Where:

              P   =    fraction (P for pesticides/Aroclors)

              LCS =    indicates a Laboratory  Control  Sample

              ##  =    suffix consisting of characters or numbers or both that
                       makes  the EPA sample number for the  LCS  unique in the
                       SDG.

               (1)  =    When reporting results  on Form  I, a  "(1)" is appended on
                       to the sample number to indicate  that the results are
                       from Gas Chromatograph  (GC) column(l)  [e.g.,  PLCSOl(l)  ].

               (2)  =    When reporting results  on Form  I, a  "(2)" is appended on
                       to the sample number to indicate  that the results are
                       from GC column(2)  [e.g.,  PLCSOK2) ].

3.3.7.5      Volatile and semivolatile standards shall  be  identified as
              FSTD***##.

              Where:

              F   =     Fraction code  (V for volatiles;  S for semivolatiles)

              STD =     Standard    (

              *** =     Concentration of volatile standards in  ug/L (e.g., 0.5,
                        001,  005,  010,  and 025)  or the amount injected in ng for
                        semivolatile standards (e.g.,  005,  010, 020, 050, and
                        080)

              ##  =     One or two characters, numbers,  or  combinations of both
                        to create  a unique EPA sample  number within an SDG.

3.3.7.6       The Contractor  shall use the following scheme to  identify
              pesticide/Aroclor standards:

                                                       EPA Sample Number

               Individual Mix A (low point)                  INDAL##
               Individual Mix A (mid-point)                  INDAM##
               Individual Mix A (high point)                 INDAH##
               Individual Mix B (low point)                  INDBL##
               Individual Mix B (mid-point)                  INDBMi#
               Individual Mix B (high point)         '        INDBH##
               Resolution Check                              RESC##
               Performance Evaluation Mixture                 PEM##
               Toxaphene                                    TOXAPH##
               Aroclor 1016                                 AR1016##
               Aroclor 1221                                 AR1221##
               Aroclor 1232                                 AR1232##
               Aroclor 1242                                 AR1242##
               Aroclor 1248                                 AR1248##
                                           B-31                                    OLC03.2

-------
Exhibit B -- Section 3
Form Instructions (Con't)


                               Name                     EPA Sample Number
              Aroclor 1254                                 AR1254##
              Aroclor .1260                                 AR1260##
              Aroclor 1016/1260 Mix                        AR1660##

              The Contractor shall replace the two-character terminator (##)  of
              the identifier with one or two characters or numbers,  or a
              combination of both, to create a unique EPA sample number within
              an SDG.

3.3.7.6.1        If the  standards are injected onto both GC columns on the same
                 instrument  simultaneously, the same EPA sample number may be
                 used for reporting data for the standards for both columns.   If
                 simultaneous  injections are not made, then the same number
                 shall not be  used.

3.3.7.7       The EPA sample number for florisil  shall be FLO###«#####,  where
              ######### is the florisil cartridge lot number.   If the florisil
              cartridge lot number is more than nine characters,  truncate at the
              ninth character.

3.3.8     Other Common Fields.  Several other pieces of information are common
          to many of the data  reporting forms.  These  include purge
          volume/sample  volume, lab  sample  identifier, lab file  identifier,
          instrument ID, and page _  of _.

3.3.8.1       "Purge Volume" or "Sample Volume" is the total volume of water
              that was purged or extracted,  in milliliters (mL).
      P.
3.3.8.'2       The lab sample identifier is a unique laboratory-generated
     -,i       internal identifier pertaining to a particular analysis.  The
              Contractor can enter up to 12 alpha-numeric characters in the "Lab
              Sample ID"  field.  The Contractor may use the EPA sample number as
      -•"       the lab sample identifier.

3.3.8.3       The lab file identifier is the unique laboratory-generated name of
              the GC/MS data system file containing information pertaining to a
              particular analysis.  The Contractor can enter up to 14
              alpha-numeric characters in the "Lab File ID" field.

3.3.8.4       The "Instrument ID"  field is common to the forms containing
              calibration data.  The identifier used by the Contractor shall
              include some indication of the manufacturer and/or model of the
              instrument, and shall contain additional characters that
              differentiate between all instruments of the same type in the
              laboratory.

3.3.8.5       The GC column identifier,  and inner diameter are common to many of
              the reporting forms for the volatile and pesticide fractions.  In
              addition,  column length is entered  on the volatile reporting
              forms.  Under "GC Column",  enter the column identification as
              denoted by the manufacturer.  Enter the inner diameter in the "ID"
              field in millimeters (mm)  (to two decimal places),  and the column
              length in the "Length" field in m (in whole numbers).

3.3.8.6       Forms II,  III, IV, V,  VIII,  IX,  and X contain a field labeled
              "page _ of _" in the bottom lefthand corner.  If the number of
              entries required on any of these forms exceeds the available
              space, continue entries on another  copy of the same
              fraction-specific form, duplicating all header information.   If a
              second page is required, number the pages consecutively (i.e.,

OLC03.2                                    B-32

-------
                                                                   Exhibit B -- Section 3
                                                                        Form Instructions
                                                                                  Form Is
              "page  1 of  2" and  "page  2 of 2").   If a second page  is not
              required, number the page "page  1 of 1."

              NOTE:  These forms  are  fraction-specific.  For'example, Form II  LCV
              and  Form  II LCSV are for different  data.  Therefore, do  not number
              the  pages of all three versions  of  Form II as "1 of  6,"  "2  of 6,"
              etc.   Number only  pages  corresponding to  the fraction-specific
              form.

3.3.9     Rounding Rule.   For rounding off numbers to the appropriate level of
          precision, the  Contractor shall follow these rules.  If the figure
          following those to be retained is less than 5,  drop it  (round down).
          If  the figure  is greater than or equal to 5,  drop it and increase the
          last digit to be retained by 1 (round up).

3.4    Organic Analysis Data  Sheet  (Form  I,  All  Fractions)

3.4.1     Purpose.  This  form is used for tabulating and reporting sample
          analysis, including dilutions,  re-analysis, blank,  Laboratory  Control
          Sample for target compounds and requested MS/MSD. .If all fractions
          are not requested for analysis, only the pages for the  fractions
          required  shall  be submitted.  For example,  if only volatiles analysis
          is  requested,  Form I LCV-1, LCV-2 and  Form I LCV-TIC shall be
          submitted.   If  only the pesticide/Aroclor fraction is requested for
          analysis. Form I LCP  shall  be submitted.  Furthermore,  pesticide
          instrument blanks  (PIBLKs)  shall be reported on a per column/per
          analysis  basis  on Form I  LCP.  Each PIBLK shall be named with  a
          unique EPA sample number.   Also, the Laboratory Control Sample and
          the MS/MSD shall be reported on a per  column basis.  Distinguish
          between GC column(1)  and  GC column (2)  results by appending  a- suffix
          n(l)n for GC column(l) and  "(2)" for GC column(2).

3.4.2     Instructions.   Complete the header  information according to  the
          instructions in Section 3.3.  Complete the remainder of the  form
          using the following instructions.

3.4.2.1       Enter the date  of  sample receipt at the laboratory,  as noted on
              the TR (i.e.,  the  VTSR), in the "Date Received"  field.   The date
              shall be entered as MM/DD/YYYY.

3.4.2.2       Complete the "Date Extracted" and "Date Analyzed"  fields in the
              same format (MM/DD/YYYY).   For the continuous liquid-liquid
              extraction procedures, enter the date  that  the  procedure was
              started in the "Date Extracted" field.   If  separatory funnel
              (pesticides only)  was used,  enter the  date  the  procedure was
              completed in the "Date Extracted"  field.   For pesticide/Aroclor
              samples,  enter the date of  the first GC analysis performed in the
              "Date Analyzed" field.  The date of sample  receipt will be
              compared with the extraction and analysis dates of each fraction
              to ensure that contract holding times  were  not  exceeded.

3.4.2.3       For volatiles on Form I LCV-1 and LCV-2,  enter the GC column
              identifier  in the "GC Column" field, the internal diameter in mm,
              to two decimal places in the "ID"  field,  and the length in meters
              (m), as a whole number, as described in Section 3.3.

3.4.2.4       For pesticides/Aroclors, enter  the method .of extraction in the
              "Extraction" field on Form I LCP as "SEPF"  for separatory  funnel,
              or "CONT"  for continuous liquid-liquid extraction.
                                           B-33                                    OLC03.2

-------
Exhibit B -- Section 3
Form Instructions
Form Is (Con't)

3.4.2.5       For semivolatiles and pesticides/Aroclors,  enter the actual volume
              of'the most concentrated sample extract,  in microliters (uL) ,  in
              the "Concentrated Extract Volume" field on Form I LCSV-1,  LCSV-2
              or LCP.   For semivolatiles,  this volume will typically be 1,000
              uL.  For pesticides/Aroclors,  the volume of the most concentrated
              extract will typically be 2,000 uL.   If a dilution of the sample
              extract is made in a subsequent analysis,  this volume will remain
              the same,  but the dilution factor will change.

3.4.2.6       For semivolatiles and pesticides/Aroclors,  enter the volume of the
              sample extract injected into the GC in the "Injection Volume"
              field on Form I LCSV-1, LCSV-2 or LCP.  Report this volume in
              microliters (uL)  to one decimal place (e.g., 1.0 uL) .

3.4.2.7       If pesticides/Aroclors are analyzed using two GC columns connected
              to a single injection port,  enter the amount of half the volume in
              the syringe in the "Injection Volume" field (i.e., assume that .the
              extract injected is evenly divided between the two columns).

3.4.2.8       If a sample or sample extract has been diluted for analysis, enter
              the dilution factor as a single number (e.g.,  enter 100.0 for  a 1
              to 100 dilution of the sample) in the "Dilution Factor" field.
              The dilution factor shall not be entered as a fraction.  If a
              sample was not diluted, enter 1.0.  Report dilution factors to one
              decimal place.

3.4.2.9       If sulfur cleanup is employed, enter Y in the "Sulfur Cleanup"
              field; if not, enter N on Form I LCP.

3.4.2.10      For positively identified target compounds, the Contractor shall
              report the concentrations as uncorrected for blank contaminants.

3.4.2.:il      Report all analytical results to two significant figures.

3.4.2.12      Under the column labeled "Q" for qualifier, flag each result with
              the specific data reporting qualifiers listed below.  When
              reporting results to USEPA,  the Contractor shall use these
              contract-specific qualifiers.  The Contractor shall not modify the
              qualifiers.  Up to five qualifiers may be reported on Form I for
              each compound.  The Contractor is encouraged to use additional
              flags or footnotes (see the X qualifier).

              The USEPA-defined qualifiers to be used are:-

              U: This flag indicates the compound was analyzed for but not
              detected.  The Contract Required Quantitation Limit  (CRQL) shall
              be adjusted according to the equation listed in Exhibit D.  CRQLs
              are listed in Exhibit C.

              J: This flag indicates an estimated value.   This flag is used  (1)
              when estimating a concentration for tentatively identified
              compounds where a 1:1 response is assumed,   (2) when the mass
              spectral and retention time data indicate the presence of a
              compound that meets the volatile and semivolatile GC/MS
              identification criteria, and the result is less than the CRQL  but
              greater than zero, and  (3) when the retention time data indicate
              the presence of a compound that meets the identification criteria
              for a pesticide and/or an Aroclor, and the result is less than the
              CRQL but greater than zero.   For example, if the sample
              quantitation limit is 5.0 ug/L, but a concentration of 3.0 ug/L is
              calculated, report it as 3.0J.


OLC03.2                                    B-34

-------
                                                     Exhibit B -- Section  3
                                                          Form Instructions
                                                            Form Is  (Con't)

NOTE: The J flag is not used and the compound is not reported as
being identified for pesticide/Aroclor results less than the CRQL
if the pesticide residue analysis expert determines that the peaks
used for compound identification resulted from instrument noise or
other interferences (column bleed, solvent contamination, etc.).

N: This flag indicates presumptive evidence of a compound.  This
flag is only used for Tentatively Identified Compounds (TICs),
where the identification is based on a mass spectral library
search.  It is applied to all TIC results.  For generic
characterization of a TIC, such as chlorinated hydrocarbon,  the N
flag is not used.

P: This flag is used for a pesticide/Aroclor target analyte when
there is greater than 25% difference for detected concentrations
between the two GC columns (see Form X).  The lower of the two
values is reported on Form I and flagged with a P.  The P flag is
not used unless a compound is identified on both columns.

C: This flag is not used under this contract, but it is reserved
for USEPA use.

B: This flag is used when the analyte is found in the associated
blank as well as in the sample.  It indicates probable blank
contamination and warns the data user to take appropriate action.
This flag shall be used for a TIC as well as for a positively
identified target compound.

The combination of flags BU or UB is expressly prohibited.  Blank
contaminants are flagged B only when they are detected in the
sample.

E: This flag identifies compounds whose concentrations exceed the
upper  level of the calibration range of the instrument for that
specific analysis.  If one or more compounds have a concentration
greater than the upper level of the calibration range, the sample
or extract shall be diluted and re-analyzed according to the
specifications in Exhibit D; exceptions are also noted in Exhibit
D.  All such compounds with concentrations greater than  the upper
level  of the calibration  range shall have the concentrations
flagged with an E on Form I for the original analysis.   The
results of both analyses  shall be reported on separate copies of
Form I.  The Form I for the diluted sample shall have for the
volatile and semivolatile dilutions "DL"  or  ("DL2", when this
additional dilution was approved by the Region) and for  the
pesticides dilution "DL"  or "DL2"  (or  "DL3", when approved by the
Region) suffix appended to the sample number.

NOTE:  For total xylenes,  where three isomers are quantified as two
peaks, the calibration range of each peak shall be considered
separately.  For example, a diluted analysis is not required for
total  xylenes unless the  concentration  of the peak representing
the  single Isomer exceeds 25 ug/L or the  peak representing the two
co-eluting isomers on that GC column exceeds 50 ug/L.

D: This flag  is used for  all compounds  identified  in  an  analysis
as diluted.   If  a sample  or extract is  re-analyzed with  a
dilution  factor  greater than 1,  for example, when the
concentration of the analyte exceeds the  upper  calibration range,
the  "DL",  "DL2"  or  "DL3"  suffix  is appended  to  the  sample number
on Form  I  for the more diluted sample,  and all  reported


                             B-35                                    OLC03.2

-------
Exhibit B -- Section 3
Form Instructions
Form I LCV-TIC and LCSV-TIC

              concentrations on that Form I are flagged with the D flag.  This
              flag alerts data users that any discrepancies between the reported
              concentrations may be due to dilution of the sample or extract.

              NOTE 1:  The D flag is not applied to compounds which are not
              detected in the sample analysis,  i.e.,  compounds reported with the
              CRQL and the U flag.

              NOTE 2:  Separate Form Is are required for reporting the original
              analysis (EPA Sample No. XXXXX)  and the more diluted sample
              analyses, i.e., the results from these analyses cannot be combined
              on a single Form I.

              A:  This flag indicates that a Tentatively Identified Compound
              (TIC) is a suspected aldol-condensation product.

              X:Other specific flags may be required to properly define the
              results.  If used, the flags shall be fully described in the SDG
              Narrative.  Begin by using X.  If more than one flag is required,
              use Y and Z as needed.  If more than five qualifiers are required
              for a sample result, use the X flag to represent a combination of
              several flags.  For instance, the X flag might combine the B and D
              flags for some samples.  The laboratory-defined flags are limited
              to X, Y, and Z.

3.5    Organic Analysis Data Sheet: Tentatively Identified Compounds
      ; (Form I LCV-TIC and Form  I LCSV-TIC)

3.5.1:'i's    Purpose.  These forms are used to report  analysis  results for non-
     :i'•-•    target compounds (e.g.,  compounds not  listed in Exhibit C) ,  excluding
           deuterated  monitoring compounds and  internal standards.   See Exhibit
           D for  instructions on identification and  quantitation.   The
     ••'iP    Contractor  shall submit Form I LCV-TIC or LCSV-TIC for every
     •aii    analysis,  including required dilutions and re-analyses,  and blanks,
           even  if no  TICs are found.
      .s
3.5.2      Instructions.   Complete the header information according to the
           instructions in Section 3.3.   Complete the remainder of the form
           using  the  following instructions in  addition to the instructions in
           Section 3.4.

3.5.2.1       Report all TICs including CAS number (if applicable), compound
              name, retention time, and the estimated concentration as
              uncorrected for blank contaminants.  Report all analytical results'
              to two significant figures.   (Criteria for reporting TICs are
              given in Exhibit D, Section 11).   Retention time shall be reported
              in minutes and decimal minutes, not seconds or minutes:seconds.

3.5.2.2       Total the number of TICs found, and enter this number in the
              "Number TICs found" field.  If no TICs were found, enter 0  (zero).

3.5.2.3       If the name of a compound exceeds the 28 spaces in the TIC column,
              truncate the name  to 28 characters.  If the compound is an
              unknown, restrict  the description to no more than 28 characters
              (e.g.,  unknown hydrocarbon).
OLC03.2                                    B-36

-------
                                                                   Exhibit B -- Section 3
                                                                        Form Instructions
                                                                   Form II LCVs and LCSVs
3.6    Deuterated Monitoring Compound (DMC)  Recovery (Form II  LCV-1,  LCV-2  and
       Form II LCSV-1,  LCSV-2)

3.6.1     Purpose.  For volatiles and semivolatiles, Form II LCV-1, LCV-2 and
          Form II LCSV-1, LCSV-2 are used to report the recoveries of the DMCs
          added to each volatile and semivolatile sample, including dilutions,
          re-analyses, blanks and requested MS/MSD.  The DMCs are used to
          monitor the performance of the purge and trap for volatiles, the
          extraction and injection for semivolatiles, and the GC/MS system as a
          whole.

3.6.2     Instructions.  Complete the header information according to the
          instructions in Section 3.3.

3.6.2.1       For each volatile DMC  listed in Table 2 and each semivolatile DMC
              listed in Table 3,  report  the percent recovery to the nearest
              whole percentage point,  and to the  number of significant figures
              given by the QC limits at  the bottom of the form.
                                         Table 2
                         Volatile Deuterated Monitoring  Compounds
Volatile Deuterated
Monitoring Compounds
VDMC1
VDMC2
VDMC3
VDMC4
VDMC5
VDMC6
VDMC7
VDMC8
VDMC9
VDMC10
VDMC11
VDMC12
VDMC13
VDMC14
Vinyl Chloride-d3
Chloroethane-d5
1 , 1 - Dichl oroethene - d2
2 -Butanone-d5
Chloroform-d
1 , 2 -Dichloroethane-d4
Benzene-d6
1, 2-Dichloropropane-d6
Toluene -d8
trans-l,3-Dichloropropene-d4
2-Hexanone-d5
Bromoform-d
1, l,2,2-Tetrachloroethane-d2
1, 2-Dichlorobenzene-d4
CAS Number
6745-35-3
19199-91-8
22280-73-5
24313-50-6
865-49-6
17060-07-0
1076-43-3
93952-08-0
2037-26-5
93951-86-1
4840-82-8
2909-52-6
33685-54-0
2199-69-1
                                           B-37
                                                                                   OLC03.2

-------
Exhibit B — Section 3
Form Instructions
Form II LCP
                                         Table 3
                       Semivolatile Deuterated Monitoring  Compounds
Semivolatile Deuterated
Monitoring Compounds
SDMC1
SDMC2
SDMC2
SDMC4
SDMC5
SDMC6
SDMC7
SDMC8
SDMC9
SDMC10
SDMC11
SDMC12
SDMC13
SDMC14
SDMC15
SDMG16
Phenol -d5
bis- (2-Chloroethyl)ether-d8
2-Chlorophenol-d4
4-Methylphenol-d8
Nitrobenzene-dS
2-Nitrophenol-d4
2 , 4 -Dichlorophenol -d3
4-Chloroaniline-d4
Dimethylphthalate - d6
Acenaphthyl ene - d8
4 -Nitrophenol -d4
Fluorene-dlO
4 , 6-Dinitro-methylphenol-d2
Anthracene- dlO
Pyrene-dlO
Benzo (a) pyrene-d!2
CAS Number
4165-62-2
93952-02-4
93951-73-6
190780-66-6
4165-60-0
93951-78-1
93951-74-7
191656-33-4
85448-30-2
93951-97-4
93951-79-2
81103-79-9
93951-76-9
1719-06-8
1718-52-1
63466-71-7
3.6.2.2       Flag each DMC recovery outside the QC limits with an asterisk (*).
              The asterisk shall be placed in the last space in each appropriate
              column, under the "#" symbol.

3.6.2.3       In the "TOT OUT" column, total the number of DMC recoveries that
              were outside the QC limits for each sample.   If no DMCs were
              outside the limits,  enter 0 (zero).

3.6.2.4       If the sample is a dilution and the deuterated monitoring
              compounds (DMCs) are outside the acceptance  window,  enter the
              calculated recovery and flag the DMC recoveries with a D in the
              column under the "#" symbol.  Do not include recoveries flagged
              with a D in the total number of recoveries for each sample outside
              the QC limits.

3.6.2.5       Number all pages as described in Section 3.3.

3.7    Surrogate Recovery  (Form II LCP)

3;7.1      Purpose.   Form  II LCP is used to  report  the  recoveries  of  the
           surrogate compounds added to each pesticide/Aroclor sample,  blank.
           Laboratory Control Sample and requested  MS/MSD.
OLC03.2
                                           B-38

-------
                                                                   Exhibit B — Section 3
                                                                        Form Instructions
                                                                Form III LCV, LCSV, LCP-1

3.7.2         Instructions.   Complete the header information  according  to the
              instructions  in Section 3.3.   Complete the  remainder  of the form
              using the following instructions.

3.7.2.1       For each surrogate listed in Table 4,  report  the  percent  recovery
              to the nearest  whole percentage point.

3.7.2.2       Flag each surrogate recovery outside the QC limits  with an
              asterisk (*).   The asterisk shall  be placed in  the  last space in
              each appropriate column,  under the "#" symbol.

3.7.2.3       In the "TOT OUT" column,  total the number of  surrogate recoveries
              that were outside the QC limits for each sample.   If  no surrogates
              were outside  the limits,  enter 0 (zero).

3.7.2.4       If the samples  is a dilution and the surrogates-are outside the
              acceptance window in any analysis, enter the  calculated recovery,
              and flag the  surrogate recoveries with a D  in the column  under the
              "#" symbol.  Do not include results flagged with  a D in the total
              number of recoveries for each sample outside  the  QC limits.

3.7.2.5       The pesticide surrogate recoveries shall be reported from each GC
              column used for the analyses.  Therefore, identify each GC column
              at the top of Form II LCP, entering the stationary phase  in the
              "GC Column" field, and the internal diameter  of the column in mm
              in the "ID" field.

3.7.2.6       The assignment  of columns as "1" and  "2" is left  to the  discretion
              of the Contractor when the analyses are performed by simultaneous
              injection into a GC containing two columns.  If so analyzed, the
              assignment of "GC Column 1" and "GC Column 2" shall be consistent
              across all the reporting forms.  If the analysis  is not performed
              by simultaneous injection, then the assignment  of GC column number
              shall be based on the chronological order of  the  two analyses.

3.7.2.7       Number all pages as described in Section 3.3.

                                          Table  4
                                   Pesticide Surrogates

                Pesticide Surrogates                    CAS  Number


                Decachlorobiphenyl  (DCB)                2051-24-3
                Tetrachloro-m-xylene  (TCX)	877-09-8	

3.8    Matrix Spike/Matrix Spike Duplicate Recovery (Form  III,  All  Fractions,
       LCV,  LCSV,  LCP-1)

3.8.1     Matrix Spike/Matrix Spike Duplicate Recovery and Laboratory  Control
           Sample Recovery

3.8.1.1       Purpose.  This  form  is used  to  report the results of  the analyses
              of MS/MSD.  This  form  should only be  submitted if the analysis  of
              MS/MSD  samples  have  been requested by the Region.  Complete  Form
              III  LCP-1  for each GC  column used for analysis.

3.8.1.2       Instructions.   Complete  the  header information according to  the
              instructions in Section  3.3;   Include the EPA sample  number  for
              the  matrix spike, without the  suffixes  MS or MSD.  Complete  the
              remainder  of the  form  using  the following  instructions.  For each


                                           B-39                                    OLC03.2

-------
Exhibit B — Section 3
Form Instructions
Form III LCP-2

              Form III LCP-1 enter the Instrument ID,  the stationary phase in
              the "GC Column" field, and the internal  diameter of the column in
              mm in-the "ID" field-  The order of reporting is not important but
              must be consistent with Form X.

3.8.1.2.1         In the first table under the  "SPIKE ADDED" column, enter the
                  amount of  spike  added  in ug/L for each analyte..

3.8.1.2.2         Enter the  sample concentration  in the next column of each  spike
                  compound detected in the original sample.  If a  spike compound
                  was not detected during the analysis of the  original sample,
                  enter the  sample result as  0  (zero).

3.8.1.2.3         In the "MS CONCENTRATION" column, enter the  actual
                  concentration  of each  spike compound detected in the matrix
                  spike aliquot.

3.8.1.2.4         Calculate  the  percent  recovery  of each spike compound in the
                  matrix spike aliquot to the nearest whole percent, according  to
                  Exhibit D.  Enter the  percent recovery in the "MS % REC"
                  column.

3.8.. 1.2.5         Flag all percent recoveries outside the QC limits with  an
                  asterisk  (*).  The asterisk shall be placed  in the last space
                  of the "MS % REC" column, under the "#" symbol.

3.8.1.2.6         Follow Sections  3.8.1.2.1 through 3.8.1.2.5  to complete the
      *;••           lower table, using the results  of the analysis of the MSD
     ~&           aliquot.
     •'0-
3.8.1.-2-.7         Calculate  the  relative percent  difference  (RPD)  between the
      •St'•          matrix spike recovery  and the matrix spike duplicate recovery,
      '£'           and enter  this value in the "RPD" column.  Report the RPD  to
      «»•           the nearest whole percent.

3.8.1.2.8         Compare the RPDs to the QC  limits given on the form, and flag
                  each RPD outside the QC limits  with an asterisk  (*) in  the last
                  space of the  "RPD" column,  under the "#" symbol.

3.8.1.2.9         Summarize  the  values outside  the QC limits at the bottom of the
                  page.  No  further action is required by the  Contractor.

3.8.2     Laboratory  Control Sample Recovery  (Form III LCP-2)

3.8.2.1       Purpose.  Form III LCP-2 is used to report the results of the
              analyses of the Laboratory Control Samples.

3.8.2.2       Instructions.   Complete the header information according to the
              instructions in Section 3.3.   Complete the remainder of the form
              using the following  instructions.

3.8.2.2.1         If the LCS solution is purchased by the Contractor from a  third
                  party, report  the identification number used by  the third  party
                  to identify the  LCS lot, if available.  If the LCS solution was
                  prepared in-house, leave this entry blank.

3.8.2.2.2         The  "LCS Aliquot" is the volume in microliters  (uL) of  LCS
                  spiking solution that  was added to reagent water before
                  extraction.
OLC03.2                                    B-40

-------
                                                                   Exhibit B -- Section 3
                                                                        Form Instructions
                                                                                 Form IVs

3.8.2.2.3        The LCS is reported for each GC column.  Enter the Instrument
                 ID,' GC column, and internal diameter (ID)  for both GC columns.
                 The order of reporting is not important,  but must be consistent
                 with the information reported on Form X.   If simultaneous
                 injections are not made, the "Date Analyzed" is the earlier
                 date of the two LCS analyses.  The dates  should be entered in
                 MM/DD/YYYY format.

3.8.2.2.4        In the box (upper for Pesticides) in Form III, under "AMOUNT
                 ADDED", enter the amount in nanograms  (ng)  of each analyte
                 added to the sample.  Under "AMOUNT RECOVERED", enter the
                 amount in ng of each analyte in the sample calculated from
                 analysis.  Calculate the percent recovery of each compound in
                 the sample to the nearest whole percent,  according to Exhibit
                 D, and enter under "% REC".  Flag all percent recoveries which
                 do not meet the contract requirements with an asterisk  (*).
                 The asterisk must be placed in the last space of the percent
                 recovery column, under the "#" symbol.

3.8.2.2.5        Complete the lower box according to the instructions in Section
                 3.8.2.2.4.

3.8.2.2.6        Summarize the values outside the QC limits at the bottom of the
                 page.

                 NOTE: This means the results for both  columns.

3.9    Method Blank Summary (Form IV, All Fractions)

3.9.1      Purpose.  This  form  summarizes  the samples associated with  each
           method blank  analysis.  The Contractor  shall  submit  the appropriate
           Form IV  for each blank.

3.9.2      Instructions.   Complete  the header information according to the
           instructions  in Section  3.3.   The EPA sample  number  entered in the
           upper right-hand corner  shall  be  the same number entered on Form I
           for the  blank.  Complete  the remainder  of the form using the
           following instructions.

3.9.2.1       Complete the following fields: "Instrument ID", "Date Analyzed",
              and "Time Analyzed".   Dates shall be entered as MM/DD/YYYY.  The
              time shall be reported in military time.

3.9.2.2       Pesticide/Aroclor contaminants shall meet the identification  '
              criteria requiring analysis of the blank on two different GC
              columns (see Exhibit D PEST).  Enter the date,  time, and
              instrument ID of both analyses of the blank on the pesticide
              method blank summary  (Form IV LCP).   The information on the two
              analyses is differentiated as Date Analyzed (1),  Date Analyzed
               (2), etc.  'If the analyses were run simultaneously, the order of
              reporting is not important, but shall be consistent with the
              information reported on all other pesticide forms.  Otherwise,
              Date Analyzed (1) shall indicate the analysis on column 1, and
              Date Analyzed (2) shall indicate the analysis on column 2.

3.9.2.3       For pesticide/Aroclor blanks, enter the method of extraction as
               "SEPF" for separatory funnel, or "CONT" for continuous liquid-
              liquid extraction on Form IV LCP.

3.9.2.4       Identify the GC column, internal diameter, and length in the
              appropriate fields, as indicated in Section 3.3.


                                           B-41                                    OLC03.2

-------
Exhibit B -- Section 3
Form Instructions
Form V LCV and LCSV

3.9.2.5       For semivolatile and pesticide/Aroclor method blanks,  enter the
              date of extraction of the blank on Form IV LCSV or LCP.

3.9.2.6       If  the samples associated with pesticide/Aroclor blank are
              subjected to sulfur cleanup,  then the blank shall also be
              subjected to sulfur cleanup.   If sulfur cleanup is employed,  enter
              Y in the "Sulfur Cleanup" field; if not,  enter N on Form IV LCP.
              If  only some of the samples associated with the method blank are
              subjected to sulfur cleanup,  a separate sulfur cleanup blank is
              required (see Exhibit D PEST).   If a separate sulfur cleanup blank
              is  prepared,  complete one version of Form IV.associating all the
              samples with the method blank,  and a second version of Form IV
              listing only those samples associated with the separate sulfur
              cleanup blank.

              NOTE: Subjecting all samples associated with a method blank to
              sulfur cleanup avoids the need for two forms.

3.9.2.7       For all three fractions,  as appropriate,  summarize the samples
              associated with a given method blank in the table, entering the
              EPA sample number and lab sample identifier.  For volatiles,  enter
              the lab file identifier and the time of analysis of each sample.
              For semivolatiles, enter the lab file identifier and date of
              analysis.  For pesticides/Aroclors, enter the dates of both
              analyses as Date Analyzed (1)  and Date Analyzed (2), as discussed
              previously.

3.9.2.8       Number all pages as described in Section 3.3.

3.10  ;;GC/MS Instrument Performance Check (Form V LCV and Form V LCSV)

3.10.1    Purpose.   This form is used  to report  the  results  of  the  GC/MS
          instrument performance check for the volatile  and  semivolatile
      ~.l'  fractions and to summarize the date and time of analyses  of  samples,
      .i   including dilutions and re-analyses,  standards, blanks, and requested
      "•!   MS/MSD associated with each  analysis of the  instrument  performance
          check  solution.

3.10.2    Instructions.   Complete the  header information according  to the
          instructions in  Section 3.3.  Complete the remainder  of the form
          using  the following instructions.

3.10.2.1      Enter the date and time of injection of the instrument performance
              check solution  (BFB for volatiles--CAS Number 460004,  DFTPP for
              semivolatiles--CAS Number 5074715).  The date shall be entered as
              MM/DD/YYYY.  The time shall be reported as military time.

3.10.2.2      For volatiles, identify the GC column,  internal diameter, and
              column length on Form V LCV,  as described in Section 3.3.

3.10.2.3      For each ion listed on the form, enter the percent relative
              abundance in the right-hand column of the first table.  Report
              relative abundances to the number of significant figures given for
              each ion in the ion abundance criteria column.

              NOTE: For both BFB and DFTPP,  one or more of the high mass ions
              may exceed the abundance of the ion listed on the form as the
              nominal base peak, m/z 95 for BFB and m/z 198 for DFTPP.  Despite
              this possibility, all ion abundances shall be normalized to the
              nominal base peaks listed on Form V  (see Exhibits D).
OLC03.2                                    B-42

-------
                                                                   Exhibit B -- Section 3
                                                                        Form Instructions
                                                            Form VI LCVs, LCSVs, and LCPs

3.10.2.4     All  relative abundances  shall be  reported as a number.   If  the
             relative abundance  is  zero,  enter 0  (zero), not a  dash  or other
             non-numeric character.   Where parentheses appear,  compute the
             percentage of  the ion  abundance of the mass given  in  the
             appropriate footnote,  and enter that value in the  parentheses.

3.10.2.5     In the  lower table,  list all samples, including dilutions,  re-
             analyses, standards, blanks  and requested MS/MSD analyzed under
             that instrument  performance  check in chronological order, by time
             of analysis  (in  military time).   Refer to Section  3.3.7 for
             specific instructions  for identifying standards and blanks.

3.10.2.6     Complete the following fields  for all standards, samples,
             including dilutions, re-analyses, and blanks:  "EPA Sample No.",
              "Lab Sample ID",  "Lab  File ID",  "Date Analyzed", and  "Time
             Analyzed".

3.10.2.7     Number  all pages as described  in  Section 3.3.

3.11   GC/MS Initial  Calibration Data  (Form VI  LCV-1,  LCV-2,  LCV-3  and Form VI
       LCSV-1,  LCSV-2,  LCSV-3)

3.11.1     Purpose.  After a GC/MS system  has undergone an initial  five-point
           calibration at the  specific concentration levels described in  Exhibit
           D, and  after all initial  calibration criteria have been  met, the
           Contractor shall complete and submit this form for each  volatile or
           semivolatile target compound initial calibration performed which is
           relevant to the samples,  including dilutions, re-analyses,  and blanks
           in the  SDG, regardless of when  that.calibration was performed.

3.11.2     Instructions.  Complete the header information according to the  .
           instructions in Section 3.3.  Enter  the Case number and  SDG number
           for  the current data package, regardless of the original Case  for
           which the  initial calibration was  performed. . Complete the remainder
           of the  form using the  following instructions.

3.11.2.1      Enter the date(s)  of the calibration.   If  the  calendar date
              changes during the calibration procedure,  the  inclusive dates
              shall be recorded.   Dates shall be entered as  MM/DD/YYYY.

3.11.2.2      Enter the injection times of the first  and last of the standards
              analyzed in the "Calibration Times"  field.   Times  shall be
              reported in military time.

3.11.2.3      Enter the lab file identifier for each of  the  five calibration
              standards injected.  Complete the response factor  data for the
              five calibration points, and then calculate and report the average
              Relative Response Factor  (RRF)  for all  target  compounds.

3.11.2.4      For volatiles and semivolatiles,  report the RRFs for the
              deuterated monitoring compounds  in the calibration standards.  The
              Contractor shall report  the Relative Standard Deviation (%RSD) for
              all compounds.  See Exhibit D for equations.

3.12    GC/EC Initial Calibration Data (Form VI LCP-1, LCP-2, LCP-3)

3.12.1     Purpose.  The  initial  calibration of pesticides/Aroclors involves the
           determination  of  retention times,  retention time  windows,  and
           calibration factors.   For single  component  pesticide  target
           compounds,  these  data  are calculated from  the  analyses of  the
           Individual Standard Mixtures A and B at three  different  concentration


                                           B-43                                    OLC03.2

-------
Exhibit B — Section 3
Form Instructions
Form VI LCPs (Con't)

          levels.   For  the multicomponent target compounds, these data are
          calculated from a  single point calibration.

3.12.2    Instructions.  Complete one Form VI  for each GC column used for the
          three analyses of  Individual Standard Mixture A  (low-point, mid-
          point, and high-point) and the three analyses of  Individual Standard
          Mixture B during an  initial calibration.  Complete the header
          information according  to the instructions in Section 3.3.  Complete
          the  remainder of the form using the  following instructions.

3.12.2.1   •   In the "Level  (x low)"  field/  enter the concentration of  the low-
              point, mid-point, and high-point calibration standards as a
              multiplier of the low-point.   Therefore,  for the  low-point, enter
              "1.0."  The concentration of  the mid-point  standard is specified
              in Exhibit D as four times the low-point;  therefore,  enter "4.0."
              The high-point standard shall  be at least  16 times the low-point,
              but may be higher,  if that value lies within the  linear range of
              the instrument, as specified in Exhibit D.   Therefore, enter the
              appropriate multiplier to the  high-point standard concentration to
              one decimal place.

3.12.2.2      Identify the GC column and internal diameter  (in  mm)  in the
              appropriate fields.

3.12.2.3      Enter the dates of analysis of the first and last of the  six
              standards on each form in the  "Date(s)  Analyzed"  field.  Dates
              shall be entered as MM/DD/YYYY.

3.12.2.-4      For each standard analyzed,  enter the retention time of each
              applicable analyte in minutes  and decimal  minutes,  under the
       '""•'      appropriate concentration level in the "RT OF STANDARDS"  column on
       ->      Form VI LCP-1.

3.12.2/5      Calculate the mean retention time of each analyte from the three
              individual mixtures,  and report it in the "MEAN RT" column on Form
              VI LCP-1.

3.12.2.6      Calculate the retention time window for each analyte using the
              specifications in Exhibit D,  and enter the lower  limit of the
              window in the  "RT WINDOW" column under "FROM," and the upper limit
              of the window under "TO" on Form VI LCP-1.   The retention times of
              the surrogates are reported from the analyses of  Individual
              Mixture A and the windows are only required to be calculated for
              Individual Mixture A.

3.12.2.7      For the six analyses of the Individual Standard Mixtures,  the
              Contractor shall also complete the calibration factor data on Form
              VI LCP-2.   Prepare one form for each instrument and GC column
              used.  Enter the calibration factor for each compound in each of
              the standards.  Calculate and enter a mean calibration factor and
              a relative standard deviation  (%RSD).  As with surrogate retention
              times, the surrogate calibration factors are only required from
              Individual Mixture A analyses.

3.12.2.8      For the multicomponent target compounds,  the retention times,
              retention time windows,  and calibration factors shall be reported
              in a  similar fashion for each single point calibration standard.
              For each multicomponent compound, the Contractor  shall select at
              least three peaks  from each analyte,  according to the
              specifications in Exhibit D.   The retention time  and calibration
              factor data apply to each peak.  Complete one version of  Form VI


OLC03.2                                    B-44

-------
                                                                   Exhibit B -- Section 3
                                                                        Form Instructions
                                                                     Form VI LCPs (Con't)

             LCP-3  for each GC column,  for each initial calibration that
             applies to  samples  in the  data package.

3.12.3    Form VI (LCP-4) is also used to report the results of analysis of the
          Resolution Check Solution that shall begin each pesticide/Aroclor     <
          initial calibration sequence.  The Contractor shall submit one Form
          VI LCP-4 for both GC columns.

3.12.3.1     Complete the header information  as described in Section  3.3.
             Using  the same assignment  of first and  second GC  columns made for
             Form IV, enter the  GC column identifier,  internal  diameter,  and
             date and time of analysis.  Enter the EPA sample  number  for  the
             Resolution  Check Standard.  If simultaneous injections on  a  single
             GC are used, the EPA sample number may  be the same for both
             Resolution  Check Standards. If  simultaneous injections  are  not
             used,  use different suffixes to  identify the standards.  Complete
             the remainder of the form using  the  following instructions.

3.12.3.2     List each analyte,  in retention  time order, including both
             surrogate compounds.  Thus, the  order of analytes in the two boxes
             on this  form will be different due to the dissimilarity  of the
             stationary  phases of the two GC  columns used.   Enter the name of
             each target analyte in  the Resolution Check Mixture as  it  appears
             on Form  I LCP.  Spell out the names  of  the surrogates as they
             appear on Form  II LCP-2.

3.12.3.3     Enter  the retention time of each analyte from  the analysis in the
              11RT" column.

3.12.3.4     Calculate the resolution between each pair of  analytes.  Enter the
              resolution  between  the  first  and second peaks  on the line  for the
              first  analyte  listed in the box.  Enter the resolution  between the
              second and  third peaks  on the  line for  the second analyte, and so
              on, until  the  resolutions of  all possible pairs of adjacent
              analytes have been  entered.

             NOTE:  Only eight  of the nine  resolution fields  will be  filled.

3.12.4     Form VI (LCP-5, LCP-6  and  LCP-7 for each PEM, initial mid-level
           calibration mixture A,  and initial  mid-level calibration mixture B,
           respectively)  shall be used to report the  percent resolution  between
           each pair of analytes  according to  the  definition in Exhibit  D
           Pesticides and Aroclors.

           NOTE: These forms shall also be used to report all percent  resolution
           data for  the PEM and midpoint concentration Individual Mixtures A and
           B analyzed as  part of  calibration verification  (Exhibit D/PEST,
           Section 9.3).

3.12.4.1      Complete the header information as described in Section 3.3.
              Using the  same assignment  of first and second GC columns made for
              Form IV,  enter the GC column identifier, internal diameter, and
              date and time of analysis.  Enter the EPA sample number for the
              respective standards.    If  simultaneous injections are not used,
              use different suffixes  to  identify the standards.  Complete the
              remainder  of the form using the following instructions.

3.12.4.2      List each  analyte,  in retention time order,  including both
              surrogate  compounds.  Thus, the order of analytes in the two boxes
              on this form will be different due to the dissimilarity of the
              stationary phases of the two GC columns used.   Enter the name of


                                           B-45                                    OLC03.2

-------
Exhibit B — Section 3
Form Instructions
Form VII LCVs, LCSVs and LCPs

              each target analyte in the standard as it appears on Form I LCP.
              Spell out the names of the surrogates as they appear on Form II
              LCP-2.

3.12.4.3      Enter the retention time of each analyte from the analysis in the
              "RT" column.

3.12.4.4      Calculate the resolution between each pair of analytes.  Enter the
              resolution between the first and second peaks on the line for the
              first analyte listed in the box.  Enter the resolution between the
              second and 'third peaks on the line for the second analyte, and so
              on,  until the resolutions of all possible pairs of adjacent
              analytes have been entered.

              NOTE: The last resolution field will be left blank in each table.

3.13   GC/MS Continuing Calibration Data  (Form VII LCV-1, LCV-2, LCV-3 and Form
       VII LCSV-1, LCSV-2,  LCSV-3)

3.13.1    Purpose.   For volatiles and semivolatiles,  this  form is  used to
          report  the calibration of  the GC/MS system by the  analysis of
          specific calibration standards.   Form VII  is required for each 12-
          hour time  period for both  volatile and semivolatile  target compound
          analyses.   The Contractor  shall analyze calibration  standards and
          meet all criteria outlined in Exhibit D for the  minimum  Relative
          Response Factors (RRF)and  maximum percent  difference between initial
          and continuing calibrations.

3.13.2    Instructions.  Complete the header information according to the
      '«'?*  instructions in  Section 3.3.  .Complete the remainder of  the form
      -'•'  using the  following instructions.

3.13.2-:'-i      Enter the date and time of the continuing calibration and the
      -'•"-•      dates and times of the initial calibration (give inclusive dates
              if  the initial calibration is performed over more than one date).
              Dates shall be entered as MM/DD/YYYY.  Times shall be reported in
              military time.

3.13.2.2      Using the appropriate initial calibration  (volatile or
              semivolatile), enter the average RRF for each target compound and
              each deuterated monitoring compound for volatiles and
              semivolatiles.

3.13.2.3      Report the RRF  (RRF5 for Volatiles and RRF20 for Semivolatiles)
              from the continuing calibration standard analysis.

3.13.2.4      Calculate the percent difference  (%D) for all compounds.  See
              Exhibit D for equation.  If the %D is greater than 999.9, report
              as  999.9.  If the %D is less than -99.9, report as -99.9.

3.14   GC/ECD Calibration Verification Summary  (Form VII, LCP-1, LCP-2)

3.14.1    Purpose.  Form VII is used to report the results of  the  Performance
          Evaluation Mixtures (PEMs)  and the mid-point concentrations of
          Individual Standard Mixtures A and B  that,  along  with the PEM,
          bracket each 12-hour period of sample analyses.  The Contractor  shall
          submit  Form VII  LCP-1 for  each 12-hour sequence  analyzed.   Form VII
          LCP-2 shall be completed each time the Individual  Standard Mixtures
          are analyzed,  for each GC  column used.
OLC03.2                                    B-46

-------
                                                                   Exhibit B -- Section 3
                                                                        Form Instructions
                                                                  Form VIII LCV and LCSVs

3.14.2    Instructions.  Complete Form VII LCP-1 and LCP-2 for each standard
          reported on Form VIII LCP.  Complete the header information according
          to the instructions in Section 3.3.  Complete the remainder of the
          form using the following instructions.

             FORM VII  LCP-1

3.14.2.1     Enter the date(s)  of  the  initial  calibration(s).  Give  inclusive
             dates if  the  initial  calibration  is performed over more than one
             day.  Dates shall  be  entered as MM/DD/YYYY.

3.14.2.2     Identify  the  GC column and internal diameter in the  appropriate
             fields.

3.14.2.3     On Form VII LCP-1,  enter  the EPA  sample number,  lab  sample
             identifier and  date and time of analysis for the instrument blank
             that preceded the  12-hour sequence (PIBLK).  For the PEM that
             initiated or  terminated the 12-hour  sequence (PEM),  enter the EPA
             sample  number,  lab sample identifier,  and date  and time of
             analysis.

3.14.2.4     When reporting  data for the PEM at the beginning of  the initial
             calibration sequence, leave the "EPA Sample  No.",  "Lab  Sample ID",
              "Date Analyzed", and "Time Analyzed"  fields  blank for the
             instrument blank (PIBLK), when no instrument blank is analyzed
             before  the PEM.  When reporting all  other PEM  analyses, the
             instrument blank fields shall be  completed.

3.14.2.5     In the  table, report the  retention time for  each analyte in the
             PEM as  well as  the retention time'windows.

3.14.2.6     For each  analyte in the PEM, enter the amount  of the analyte found
              in the  PEM,  in  ng to three decimal places,  in  the "CALC AMOUNT"
             column.

3.14.2.7     - Enter  the nominal amount   (amount  added) of each analyte in the PEM
              in ng  to  three  decimal places in  the "NOM AMOUNT" column.

3.14.2.8      Calculate the percent difference  between the calculated amount and
             nominal amount  for each analyte according to Exhibit D.  Report
              the values in the "%D" column.   If the %D is greater than 999.9,
              report  as 999.9.  If the %D is less  than -99.9, report  as -99.9.
                               1
3.14.2.9      Calculate the percent breakdown for  endrin and 4,4'-DDT and the
              combined percent breakdown in the PEM according to Exhibit D.
              Enter the values for the breakdown of endrin and 4,4"-DDT in their
              respective fields immediately under  the table.

              FORM VII  LCP-2

3.14.2.10     The upper table on Form VII LCP-2 contains the retention time and
              amount data for Individual Standard Mixture A compounds.  The
              lower table contains the data for Mixture B.  Complete the form
              using the instructions in Sections 3.14.2.1 through 3.14.2.8 for
              Form VII LCP-1.

3.15    Internal  Standard Area and  RT Summary (Form VIII LCV and Form VIII LCSV-
        1, LCSV-2)

3.15.1    Purpose.   This form is used  to summarize the peak  areas and retention
           times of the internal standards added to the initial calibration


                                           B-47                                    OLC03.2

-------
Exhibit B — Section 3
Form Instructions
Form VIII LCV and LCSVs  (Con't)

          standards, continuing calibration standards and all volatile and
          semivolatile samples, including dilutions, re-analyses, and blanks.
          The data are used to determine when changes in internal standard
          responses will adversely affect quantitation of target compounds.
          This  form shall be completed each time an initial calibration and
          continuing calibration is performed, or when samples are analyzed
          under the same GC/MS instrument performance check as an initial
          calibration.

3.15.2    Instructions.  Complete the header information according to Section
          3.3.   Complete the remainder of the form using the following
          instructions.  If samples are analyzed immediately following an
          initial calibration, before another instrument performance check and
          a continuing calibration, Form VIII shall be completed on the basis
          of the internal standard areas of the 5 ug/L initial calibration
          standard for volatiles, and the 20 ng initial calibration standard
          for semivolatiles.   Use the date and time of analysis of this
          standard and the lab file identifier and areas in place of those of a
          continuing calibration standard.

3.15.2.1      Enter the date and time  of  analysis  of the continuing calibration
              standard.   The date shall be entered as MM/DD/YYYY.   The time
              shall be reported in military time.

3.15.2.2      For volatiles,  enter the GC column identifier,  internal  diameter,
              and length as  directed in Section 3.3.

3.15.2.3      From the results of the  analysis  of  the continuing calibration
      ••''?''      standard,  enter the area measured for each internal  standard and
      ••'•"       its retention time (in decimal minutes)  under the appropriate
              column in the  "12 HOUR STD" row.
      t: •«.;*•
3.15.2-/4      For each internal standard  listed in Tables 5 and 6,  calculate the
              upper limit of the area  and the lower limit of the area from the
              internal standard area.   For FORM VIII LCV,  calculate the upper
              limit of the area as the area of  the particular internal standard
              plus 40 percent of its area (i.e.,  1.4 times the area in the "12
              HOUR STD" field), and the  lower limit of the area as the area of
              the internal standard minus 40 percent of its area (i.e.,  0.6
              times the area in the "12 HOUR STD"  field).   For FORMS VIII  LCSV-1
              and VIII LCSV-2,  calculate  the upper limit of the area as the area
              of the particular internal  standard  plus 100 percent of  its  area
              (i.e.,  two times the area  in the  "12 HOUR STD" field),  and the
              lower limit of the area  as  the area  of the internal  standard minus
              50 percent of  its area (i.e.,  0.5 times the area in the  "12  HOUR
              STD" field).   Report these  values in the "UPPER LIMIT"  and "LOWER
              LIMIT"  rows,  respectively.   Calculate the upper limit of the
              retention time and the lower limit of the retention time.  The
              upper limit of the retention time is calculated by adding 0.33
              minutes to the retention time of  the internal standard and the
              lower limit of the retention time is the retention time  of the
              internal standard minus  0.33 minutes.   Report these values in the
              "UPPER LIMIT"  and "LOWER LIMIT" rows in the applicable RT columns.

3.15.2.5      For each sample,  including  dilutions,  re-analyses,  blanks,  and
              requested MS/MSD analyzed under a given continuing calibration,
              enter the EPA sample number and the  area measured for each
              internal standard and its  retention  time.   If the internal
              standard area  is outside the upper or lower limits calculated in
              Section 3.15.2.4, flag that area  with an asterisk (*).   The
              asterisk shall be placed in the far  right-hand space of  the  box


OLC03.2                                    B-48

-------
                                                                   Exhibit B -- Section 3
                                                                        Form Instructions
                                                                            Form VIII LCP
              for each  internal standard area, directly under the  "#"  symbol.
              Similarly,  flag  the  retention  time of any internal standard that
              is outside  the limits with an  asterisk.

3.15.2.6      Number  all  pages as  described  in Section 3.3.

                                         Table 5
                               Volatile Internal Standards


         Volatile Internal Standards	CAS Number	

         IS1: Chlorobenzene-d5 (CBZ)                          3114-55-4

         IS2: 1,4-Difluorobenzene (DFB)                       540-36-3

         IS3: l,4-Dichlorobenzene-d4 (DCB)                   3855-82-1
                                         Table 6
                             Semi-volatile Internal Standards
              Semivolatile Internal Standards                CAS Number
IS1:
IS2:
IS3:
IS4:
IS5:
IS6:
l,4-Dichlorobenzene-d4 (DCB)
Naphthalene-dS (NPT)
Acenaphthene-dlO (ANT)
Phenanthrene-dlO (PHN)
Chrysene-dl2 (CRY)
Perylene-dl2 (PRY)
3855-82-1
1146-65-2
15067-26-2
1517-22-2
1719-03-5
1520-96-3
3.16   Pesticide Analytical Sequence (Form VIII LCP)

3.16.1     Purpose.   This  form is used  to report the analytical sequence  for
           pesticide  analysis.  At  least one  form  is required for  each GC column
           used for pesticide/Aroclor analyses.

3.16.2     Instructions.   Complete  the  header information according  to the
           instructions in Section  3.3.  Complete  the  remainder of the form
           using the  following instructions.

3.16.2.1      Enter the date(s) of the initial calibration.  Give inclusive
              dates if the initial calibration is performed over more than one
              day.  Dates shall be entered as MM/DD/YYYY.

3.16.2.2      Identify the GC column and internal diameter in the appropriate
              fields.

3.16.2.3      At the top of the table,  report the mean retention time for
              tetrachloro-m-xylene  (TCX)  and decachlorobiphenyl (DCB) calculated
              from the initial calibration sequence (from INDA).

3.16.2.4      For every analysis associated with a particular analytical
              sequence starting with the initial calibration,  enter the EPA
              sample number, lab sample identifier, and date and time of
              analysis.  Each sample analyzed as part of the sequence shall be
              reported on Form VIII LCP even if it  is not associated with the
              SDG.  The Contractor shall use ZZZZZ  as the EPA sample number to-
                                           B-49                                    OLC03.2

-------
Exhibit B -- Section 3
Form Instructions
Form IX LCP

              distinguish all samples that are not part of the SDG being
              reported.

3.16.2.5      Report the retention time of the surrogates for each analysis in
              the "TCX RT" and "DCB RT" columns.   All sample analyses shall be
              bracketed by acceptable analyses of instrument blanks,  a PEM, and
              Individual Standard Mixtures A and B.  Given the fact that the
              initial calibration may remain valid for some time (Exhibit D),  it
              is only necessary to report the data from 12-hour periods when
              samples,  dilutions, re-analyses, Laboratory Control Samples,
              requested MS/MSD, blanks, or multicomponent standard analytes for
              the 72-hour confirmation requirement in an SDG were analyzed.  The
              Contractor shall submit Form VIII for the initial calibration
              sequence and forms that include the PEMs and Individual Standard
              Mixtures that bracket any and all samples in the SDG.  While the
              data for time periods between the initial calibration and samples
              in the SDG are not a routine deliverable, the data shall be
              available as requested (e.g., at on-site evaluations).   Non-USEPA
              samples shall be numbered ZZZZZ.

3.16.2.6      Flag all those values which do not meet the contract requirements
              by entering an asterisk  (*) in the "RT" column, under the "#"
              symbol.  If the retention time cannot be calculated due to
              interfering peaks, leave the "RT" column blank for that surrogate,
              enter an asterisk in the last column, and document the problem in
              the SDG Narrative.

3.16.2:7      If more than a single copy of Form VIII LCP is required, enter the
      •'••      same header information on all subsequent pages for that GC column
      v.       and instrument, and number each page as described in Section 3.3.

3.17   Pesticide Cleanup Summary  (Form IX, LCP)

3.17.1^    Purpose.   Form IX LCP is used  to report  the results of  the  check of
           the Florisil  cartridges used to process  all sample extracts and  to
           associate the lot of cartridges with particular sample  results so
           that problems with a particular cartridge  lot  may be tracked across
           all associated samples.

3.17.2     Instructions.   Complete the  header information according to the
           instructions  in Section 3.3.   Enter the  Case number and SDG number
           for the  current data package,  regardless of the original Case for
           which the cartridge check was  performed.   Complete the  remainder of
           the form using the following instructions.

3.17.2.1      Enter the Florisil cartridge lot number.

3.17.2.2      Enter the date the Florisil cartridge check solution was analyzed
              in the "Date of Analysis" field.  The date shall be entered as
              MM/DD/YYYY.

3.17.2.3      Complete the "GC Column" and "ID" fields for the two GC columns
              used to analyze the samples, including blanks, re-analyses.
              Laboratory Control Samples, and requested MS/MSD.  Report all
              results from either GC Column 1 or GC Column 2.

3.17.2.4      In the first table, enter the amount of spike added and spike
              recovered in ng for each analyte.

3.17.2.5      Calculate the percent recovery to the nearest whole percent,  and
              enter the number in the "% REC" field.   Flag each spike recovery


OLC03.2                                    B-50

-------
                                                                   Exhibit B — Section 3
                                                                        Form Instructions
                                                                              Form X LCPs

              outside  the  QC limits (shown on the  form) with an  asterisk  (*).
              The  asterisk shall be placed in the  last  space in  the  "% REC"
              column,  under the "#" symbol.

3.17.2.6      In the second table,  complete  the  "EPA Sample No.,"  the  "Lab
              Sample ID,"  and "Date Analyzed" fields for  each  sample and  blank
              that were cleaned up using this lot  of Florisil  cartridges.

3.17.2.7      Number the pages as described  in Section  3.3.

3.18   Pesticide/Aroclor Identification (Form X, LCP-1, LCP-2)

3.18.1    Purpose.  This  form summarizes the quantitations of all target
          pesticides/Aroclors detected  in a given sample.  It reports the
          retention times  of  the compound on both columns on which  it was
          analyzed, as well as the retention time windows of the  standard for
          that  compound on both of these columns.   In  addition, it  is used to
          report  the  concentration determined from  each GC column,  and the
          percent difference  between  the two quantitative results.  Separate
          forms are used  for  single component analytes and multicomponent
          analytes.

          Form  X  is required  for each sample, including dilutions,  re-analyses,
          blanks.  Laboratory  Control  Samples and  requested MS/MSD  in which
          compounds listed in Exhibit C (Pesticides/Aroclors) are reported on
          Form  I.  Do not generate a  Form X for pesticide instrument blanks.

3.18.2    Instructions.   Complete the header information according  to the
          instructions in Section 3.3.   Complete  the remainder  of the form
          using the following instructions.

3.18.2.1      Enter the date(s) of analysis.  Dates shall be entered as
              MM/DD/YYYY.

3.18.2.2      Enter the GC column  and internal  diameter for each of the  two
              columns.

3.18.2.3      For each single  component pesticide positively identified,  enter
              the name of the  compound  in the "ANALYTE" column as it appears on
              Form I..

3.18.2.4      Enter the retention  times on each column of the compounds  detected
              in  the  sample next to the appropriate column designation (1 or 2).

3.18.2.5      Enter the retention  time  windows on each column from the initial
              calibration  standard.   These data shall  correspond with those on
              Form VI and  shall  be entered in a similar manner.  The lower value
              is  entered under the "FROM" column, the upper value under the "TO"
              column.

3.18.2.6      Enter the concentration calculated from  each GC column under the
              "CONCENTRATION"  column.   Although the units are the same as those
              used on Form I,  do not  enter any units on Form X.

3.18.2.7      Calculate the percent difference between the concentrations
              entered on  this form, using the equation found in Exhibit D, and
              report  it to a  tenth of a percent in  the "%D" column.  If %D is
              greater that  999.9,  report it as  999.9.

3.18.2.8      The lower of the two concentrations is reported on  Form I for each
              pesticide compound.   The  lower concentration is used because, if


                                           B-51                                    OLC03.2

-------
Exhibit B — Section 3
Form Instructions
Form DC-1

              present,  coeluting interferences are likely to increase the
              calculated concentration of any target compound.   If the percent
              difference between the calculated concentrations  is greater than
              25.0 percent,  flag the concentration on Form I, as described
              previously.   This will alert the data user to the potential
              problems in quantitating this analyte.

3.18.2.9      If more pesticide compounds are identified in an  individual sample
              than can be reported on one Form X,  complete as many additional
              copies of Form X as necessary, duplicating all header information
              and numbering the pages as described in Section 3.3.

3.18.2.10     Report multicomponent analytes detected in samples on Form X LCP-
              2.  Complete the header information and GC column fields as
              described above.  For multicomponent analytes, it is necessary to
              report the retention time and concentration of each peak chosen
              for quantitation in the target analyte in a fashion similar to-
              that for single component pesticides.  The concentrations of all
              peaks quantitated (three are required, up to five may be used)  are
              averaged to determine the mean concentration.  Report the lower of
              the two mean concentrations on Form I.  Flag this value if the
              mean concentrations from the two GC columns differ by more than 25
              percent,  as described previously.

3.18.2.11     If more multicomponent compounds are identified in an individual
              sample than can be reported on one Form X, complete as many
              additional copies of Form X as necessary, duplicating all header
       •;¥      information and numbering the pages as described  in Section 3.3.

3.19   'Sample Log-In Sheet  (Form DC-1)

3.19.lK:-  Purpose.   This  form is  used to document the  receipt  and inspection  of
           sample containers and samples.   One original of Form DC-1  is  required
           for each sample shipping container (only the hardcopy form is
       V   required).   If  the samples in a single  sample shipping container are
           assigned to  more  than one SDG,  the original  Form DC-1 shall be placed
           with the' deliverables for the SDG of the lowest alpha-numeric number;
           and a copy of Form DC-1 shall be placed with the deliverables for the
           other SDGs.   The  copies shall be identified  as "copyties)", and the
           location of  the original shall be noted on the copies.

3.19.2     Instructions

3.19.2.1      Sign and date the airbill.   (If an airbill is not received,
              include a hardcopy receipt requested from the shipping company or
              a printout of the shipping company's electronic tracking
              information).

3.19.2.2      Complete the header information on the form, including the log-in
              date.

3.19.2.3      Examine the shipping container and record the presence/absence of
              custody seals and their condition (e.g.,  intact,  broken)  in item
              1.

3.19.2.4      Record the custody seal numbers in item 2.

3.19.2.5      Open the container, remove the enclosed sample documentation,  and
              record the presence/absence of chain-of-custody record(s), SMO
              forms  (e.g., TRs, Packing Lists), and airbills or airbill stickers
              in items 3-5.  Specify if there is an airbill present or an


OLC03.2                                    B-52

-------
                                                                   Exhibit B -- Section 3
                                                                        Form Instructions
                                                                                Form DC-2
              airbill sticker in item 5.   Record the airbill or sticker number
              in item 6.

3.19.2.6      Remove the samples from the shipping container(s),  examine the
              samples and the sample tags (if present),  and record the condition
              of the sample bottles (e.g., intact,  broken,  leaking)  and presence
              or absence of sample tags in items 7 and 8.

3.19.2.7      Record the presence of the cooler temperature indicator bottle in
              item 9 and cooler temperature in item 10.

3.19.2.8      Review the sample shipping documents and compare the information
              recorded on all the documents and samples and circle the
              appropriate answer in item 11.

3.19.2.9      Record the date and time of cooler receipt at the laboratory in
              items 12 and 13.

3.19.2.10     If there are no problems observed during receipt,  sign and date
              (include the time) Form DC-1, the Chain-of-Custody record, and the
              TR, and write the sample numbers on Form DC-1 in the "EPA Sample
              #" column.

3.19.2.11     Record the appropriate sample tags and assigned laboratory
              numbers, if applicable.

3.19.2.12     Any comments should be made in the "Remarks" column.

3.19.2.13     Record the fraction designation  (if appropriate) and the specific
              area designation  (e.g., refrigerator number) in the "Sample
              Transfer" block.  Sign and date the "Sample Transfer" block.

3.19.2.14     Cross out unused columns and spaces.  Initial and date all cross
              outs.

3.19.2.15     If there are problems observed during receipt or an answer marked
              with an asterisk  (e.g.,  "absent*") was circled, contact SMO and
              document the contact as well as resolution of the problem on a CLP
              Communication Log.  Following resolution, sign and date the forms
              and note, where appropriate, the resolution of the problem.

3.20    Complete SDG  File  (CSF)  Inventory Sheet  (Form DC-2)

3.20.1     Purpose.   Form DC-2 is used to record both the CSF documents and the
           number of documents in the original Sample Data Package sent to the
           USEPA Region.

3.20.2     Instructions

3.20.2.1      Organize all USEPA  CSF documents as described in Section "2.6.
              Assemble the documents in  the order specified on Form DC-2 and
              Section  2.6, and  stamp each page with a consecutive number;
              however, do not number Form DC-2.  Inventory the CSF by reviewing
              the document numbers- and recording page number ranges in  the
              columns  provided  on Form DC-2.   The Contractor shall verify and
              record in the  "Comments" section on Form DC-2 all intentional gaps
              in the page numbering sequence  (e.g.,  "page numbers not used, XXXX
              - XXXX,  YYYY  - YYYY.  if there are no documents for a specific
              document type, enter an  "NA" in  the empty space.
                                           B-53                                    OLC03.2

-------
Exhibit_B -- Sections 3 & 4
Data Reporting Forms

3.20.2.2      Certain laboratory-specific documents related to the CSF may not
              fit into a clearly defined category.   The Contractor shall  review
              Form DC-2 to determine if it is most  appropriate to place them
              under categories 7,  8,  9, or 10.   Category 10 should be used if
              there is no appropriate previous category.   These types of
              documents should be described or listed in the blanks under each
              appropriate category on Form DC-2.

3.20.2.3      If it is necessary to insert new or inadvertently omitted
              documents, the Contractor shall identify the documents with unique
              accountable numbers and record the unique accountable numbers and
              the locations of the documents in the CSF (in the "Other Records"
              section on Form DC-2).

4.0    DATA REPORTING FORMS

       The data reporting forms are shown on the following pages.
OLC03.2                                    B-54

-------
Lab Name:

Lab Code:
Lab Sample ID:

Lab File ID: _

Purge Volume:

GC Column: 	
                                           1LCA
                   LOW CONCENTRATION WATER VOLATILE ORGANICS ANALYSIS
                                             DATA  SHEET
                                          Contract:
                       Case No.:
                   Client No.:
                                                                SDG NO.:
.(ML)
      ID:
.(MM)
                             Date Received:

                             Date Analyzed:
                                                                            EPA SAMPLE  NO.
Dilution Factor: 	

   Length:  	(M)
CAS NO.
75-71-8
74-87-3
75-01-4
74-83-9
75-00-3
75-69-4
75-35-4
76-13-1
67-64-1
75-15-0
79-20-9
75-09-2
156-60-5
1634-04-4
75-34-3
156-59-2
78-93-3
74-97-5
67-66-3
71-55-6
110-82-7
56-23-5
71-43-2
107-06-2
COMPOUND
Dichlorodifluoromethane
Chloromethane
Vinyl Chloride
Bromomethane
Chi oroethane
Trichlorofluoromethane
1, 1-Dichloroethene
1,1, 2-Trichloro-l,2, 2-trif luoroethane
Acetone
Carbon Disulfide
Methyl Acetate
Methylene Chloride
trans -1,2 -Dichloroethene
Methyl tert -Butyl Ether
1 , 1 -Dichl oroethane
cis- 1,2 -Dichloroethene
2 -Butanone
Bromochloromethane
Chloroform
1 , 1 , l-Trichloroethane
Cyclohexane
Carbon Tetrachloride
Benzene
1 , 2 -Dichloroethane
CONCENTRATION UNITS:
(UG/L)
























Q
























                                        FORM I LCV-1
                                                                                  OLC03.2

-------
Lab Name:

Lab Code:
Lab Sample ID:

Lab File ID: _

Purge Volume:

GC Column: 	
                                           1LCB
                   LOW CONCENTRATION WATER VOLATILE ORGANICS ANALYSIS
                                             DATA SHEET
                       Case No.:
.(ML)
     ID:
                                          Contract:
                   Client No.:
                                                                SDG No.:
                             Date Received:

                             Date Analyzed:
.(MM)
                                                                            EPA SAMPLE NO.
Dilution Factor: 	

   Length:  	(M)
CAS NO.
79-01-6
108-87-2
78-87-5
75-27-4
10061-01-5
108-10-1
108-88-3
10061-02-6
79-00-5
127-18-4
591-78-6
124-48-1
106-93-4
108-90-7
100-41-4
1330-20-7
100-42-5
75-25-2
98-82-8
79-34-5
541-73-1
106-46-7
95-50-1
96-12-8
120-82-1
87-61-6
COMPOUND
Trichloroethene
Me t hy 1 eye 1 ohexane
1 , 2 -Dichloropropane
Bromodichloromethane
cis-1, 3-Dichloropropene
4 -Methyl - 2 -pentanone
Toluene
trans -1, 3-Dichloropropene
1,1,2 -Trichloroethane
Tetrachloroethene
2 -Hexanone
Dibromochloromethane
1 , 2 -Dibromoethane
Chlorobenzene
Ethylbenzene
Xylene (total)
Styrene
Bromoform
Isopropylbenzene
1,1,2,2 -Tetrachloroethane
1, 3-Dichlorobenzene
i
1 , 4 -Dichlorobenzene
1, 2-Dichlorobenzene
1 , 2 -Dibromo-3 -chloropropane
1,2,4 -Trichlorobenzene
1,2,3 -Trichlorobenzene
CONCENTRATION UNITS:
(UG/L)


























Q


























                                       FORM I LCV-2
                                                                                  OLC03.2

-------
                                           1LCC
                      LOW CONCENTRATION WATER SEMIVOLATILE ORGANICS
                                       ANALYSIS DATA SHEET
                                                                           EPA SAMPLE NO.
Lab Name:

Lab Code:
               Contract:
                       Case No.:
                Client No.:
                                     SDG No.:
Lab Sample ID:

Lab File ID:
Sample Volume:
(ML)
Date Received: .

Date Extracted:

Date Analyzed:
Concentrated Extract Volume: 	

Injection Volume: 	  (UL)
             (UL)
                                                     Dilution Factor:
CAS NO.
100-52-7
108-95-2
111-44-4
95-57-8
95-48-7
108-60-1
98-86-2
106-44-5
621-64-7
67-72-1
98-95-3
78-59-1
88-75-5
105-67-9
111-91-1
120-83-2
91-20-3
106-47-8
87-68-3
105-60-2
59-50-7
91-57-6
77-47-4
88-06-2
95-95-4
92-52-4
91-58-7
88-74-4
131-11-3
606-20-2
208-96-8
99-09-2
83-32-9
COMPOUND
Benzaldehyde
Phenol
bis (2-Chloroethyl) ether
2-Chlorophenol
2 -Methylphenol
2,2' -oxybis (1-Chloropropane)
Acetophenone
4 -Methylphenol
N-Nitroso-di-n-propylamine
Hexachloroethane
Nitrobenzene
Isophorone
2 -Nitrophenol
2 , 4 -Dimethylphenol
bis (2-Chloroethoxy) methane
2 , 4 -Dichlorophenol
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
Caprolactam
4 -Chloro- 3 -methylphenol
2 -Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6-Trichlorophenol
2 , 4 , 5-Trichlorophenol
1,1' -Biphenyl
2 -Chloronaphthalene
2-Nitroaniline
Dimethylphthalate
2 , 6 -Dinitrotoluene
Acenaphthylene
3-Nitroaniline
Acenaphthene
CONCENTRATION UNITS :
(UG/L)

































Q


*






























                                       FORM I LCSV-1
                                                                                  OLC03.2

-------
                                           1LCD
                 LOW CONCENTRATION WATER SEMIVOLATILE ORGANICS ANALYSIS
                                            DATA SHEET
                                                                           EPA SAMPLE NO.
Lab Name:

Lab Code:
                                         Contract:
                       Case No.:
                Client No.:
                                                                SDG No.:
Lab Sample ID:

Lab File ID:
Sample Volume:
(ML)
Date Received:

Date Extracted:

Date Analyzed:
Concentrated Extract Volume: 	

Injection Volume: 	 (UL)
             (UL)
                                                    Dilution Factor:
CAS NO.
51-28-5
100-02-7
132-64-9
121-14-2
84-66-2
86-73-7
7005-72-3
100-01-6
534-52-1
86-30-6
95-94-3
101-55-3
118-74-1
1912-24-9
87-86-5
85-01-8
120-12-7
84-74-2
206-44-0
129-00-0
85-68-7
91-94-1
56-55-3
218-01-9
117-81-7
117-84-0
205-99-2
207-08-9
50-32-8
193-39-5
53-70-3
191-24-2
COMPOUND
2 , 4 -Dinitrophenol
4 -Nitrophenol
Dibenzofuran
2 , 4 -Dinitrotoluene
Diethylphthalate
Fluorene
4 -Chlorophenyl -phenylether
4-Nitroaniline
4, 6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine (1)
1,2,4,5 Tetrachlorobenzene
4 -Bromophenyl -phenylether
Hexachlorobenzene
Atrazine
Pentachlorophenol
Phenanthrene
Anthracene
Di -n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalat'e
3,3' -Dichlorobenzidine
Benzo (a) anthracene
Chrysene
bis (2-Ethylhexyl)phthalate
Di -n-octylphthalate
Benzo (b) f luoranthene
Benzo (k) fluoranthene
Benzo (a) pyrene
Indeno ( 1 , 2 , 3 -cd) pyrene
Dibenzo (a , h) anthracene
Benzo (g , h , i ) perylene
CONCENTRATION UNITS:
(UG/L)
































Q
































 (1) Cannot be separated from Diphenylamine
                                       FORM  I LCSV-2
                                                                                 OLC03.2

-------
                                           1LCE
                  LOW CONCENTRATION WATER PESTICIDE ORGANICS ANALYSIS
                                             DATA SHEET
                                                                           EPA SAMPLE NO.
Lab Name:

Lab Code:
               Contract:
                       Case No.:
                                           Client No.:
                                                                SDG No.:
Lab Sample ID:

Sample Volume:
(ML)
Concentrated Extract Volume: 	

Injection Volume: 	  (UL)

Sulfur Cleanup:  (Y/N) 	
             (UL)
Date Received:
Date Extracted:
Date Analyzed:
                          Dilution Factor:

                  Extraction:  (Sepf/Cont)
CAS NO.
319-84-6
319-85-7
319-86-8
58-89-9
76-44-8
309-00-2
1024-57-3
959-98-8
60-57-1
72-55-9
72-20-8
33213-65-9
72-54-8
1031-07-8
50-29-3
72-43-5
53494-70-5
7421-93-4
5103-71-9
5103-74-2
8001-35-2
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
COMPOUND
alpha-BHC
beta-BHC
delta-BHC
gamma -BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4,4' -DDE
Endrin
Endosulfan II
4,4' -ODD
Endosulfan sulfate
4,4' -DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha-Chl or dane
gamma - Chi ordane
Toxaphene
Aroclor-1016
Aroclor-1221
Aroclor-1232
Aroclor-1242
Aroclor-1248
Aroclor-1254
Aroclor-1260
CONCENTRATION UNITS:
(UG/L)




























Q




























                                         FORM I  LCP
                                                                                  OLC03.2

-------
                                           1LCF
                   LOW CONCENTRATION VJATER VOLATILE ORGANICS ANALYSIS
                           DATA SHEET TENTATIVELY  IDENTIFIED COMPOUNDS
                                                                           EPA SAMPLE NO.
Lab Name:

Lab Code:
                                         Contract:
                       Case No.:
                 Client No.:
                                                                SDG No.:
Lab Sample ID:

Lab File ID: _

Purge Volume:

GC Column: 	
                           Date Received:

                           Date Analyzed:
(ML)
    ID:
       Dilution Factor: 	

(MM)       Length:  	(M)
Number TICs found:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
CAS NT7MBER






























COMPOUND NAME






























RT






























EST. CONC.
(UG/L)






























O






























                                      FORM I LCV-TIC
                                                                                 OLC03.2

-------
                                           1LCG
                LOW  CONCENTRATION WATER SEMIVOLATILE  ORGANICS  ANALYSIS
                           DATA SHEET TENTATIVELY IDENTIFIED COMPOUNDS
                                                EPA  SAMPLE  NO.
Lab Name:

Lab Code:
                                         Contract:
                       Case No.:
                Client No.:
                                                                SDG No.:
Lab Sample ID:

Lab File ID:
Sample Volume:
(ML)
Date Received:

Date Extracted:

Date Analyzed:
Concentrated Extract Volume:

Injection Volume: 	

Number TICs found:	
              (UL)
                                                     Dilution Factor:
     (UL)

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
CAS NUMBER






























COMPOUND NAME






























RT






























EST. CONC.
(UG/L)






























0






























                                      FORM I LCSV-TIC
                                                                                  OLC03.2

-------
                                           2LCA
         LOW CONCENTRATION WATER VOLATILE DEUTERATED  MONITORING COMPOUND RECOVERY
Lab Name:

Lab Code:.
                                             Contract:
                       Case No.:
                      Client No.:
                                                                SDG No.:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























VDMC1
(VCL) #






























VDMC2
(CLA) #






























VDMC3
(DCE) #






























VDMC4
(BUT) #






























VDMC5
(CLF) #






























VDMC6
(DCA) #






























VDMC7
(BEN) #






























       VDMC1   (VCL)
       VDMC2   (CLA)
       VDMC3   (DCE)
       VDMC4   (BUT)
       VDMC5   (CLF)
       VDMC6   (DCA)
       VDMC7   (BEN)
= Vinyl Chloride-d3
= Chloroethane-d5
= l,l-Dichloroethene-d2
= 2-Butanone-d5
= Chloroform-d
= l,2-Dichloroethane-d4
= Benzene-d6
PC LIMITS
(49-138)
(60-126)
(65-130)
(42-171)
(80-123)
(78-129)
(78-121)
       #  Column to be used  to  flag  recovery values
       *  Values outside of  contract required QC limits
       Page 	 of 	
                                       FORM II LCV-1
                                                                                  OLC03.2

-------
                                           2LCB
        • LOW CONCENTRATION WATER VOLATILE DEUTERATED MONITORING  COMPOUND RECOVERY
Lab Name:

Lab Code:
                                             Contract:
                       Case No.:
Client No.:
                                                                SDG No.:

01
02
03
04
05
06
07
08
09
10
LI
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























VDMC8
(DPA) #






























VDMC9
(TOD #






























VDMC10
(TOP) #






























VDMC11
(HEX) #






























VDMC12
(BRF). #






























VDMC13
(TCA) #






























VDMC14
(DCZ) #






























TOT
OUT






























        VDMC8   (DPA)    = l,2-Dichloropropane-d6
        VDMC9   (TOL)    = Toluene-d8
        VDMC10  (TDP)    = trans-l,3-Dichloropropene-d4
        VDMC11  (HEX)    = 2-Hexanone-d5
        VDMC12  (BRF)  .  = Bromoform-d
        VDMC13  (TCA)    = l,l,2,2-Tetrachloroethane-d2
        VDMC14  (DCZ)    = l,2-Dichlorobenzene-d4
                  PC LIMITS
                  (84-123)
                  (77-120)
                  (80-128)
                  (37-169)
                  (76-135)
                  (75-131)
                  (50-150)
           Column to be used to flag recovery values
           Values outside of contract required QC limits
        Page 	 of 	
                                       FORM  II LCV-2
                                                                                   OLC03.2

-------
                                           2LCC
       LOW CONCENTRATION WATER SEMIVOLATILE OBLITERATED MONITORING COMPOUND RECOVERY
Lab Name:

Lab Code:
                                             Contract:
                       Case  No.:
Client No. -.
                                                                SDG No.:

01
02
03
04
05
06
07
08
09
LO
LI
12
13
L4
L5
L6
17
18
L9
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























SDMCl"
(PHL) #






























SDMC2
(BCE) #






























SDMC3
(2CP) #

•




























SDMC4
(4MP) #






























SDMC5'
(NBZ) #






























SDMC6
(2NP) #






























SDMC7
(DCP) #






























SDMC8
(4CA) #






























                                                                PC LIMITS
         SDMC1  (PHL)    =  Phenol-d5                             (10-110)
         SDMC2  (BCE)    =  bis-(2-Chloroethyl)ether-d8           (41-94)
         SDMC3  (2CP)    =  2-Chlorophenol-d4                     (33-110)
         SDMC4  (4MP)    =  4-Methylphenol-d8                     (38-95)
         SDMC5  (NBZ)    =  Nitrobenzene-d5                       (35-114)
         SDMC6  (2NP)    =  2-Nitrophenol-d4                       (40-106)
         SDMC7  (DCP)    =  2,4-Dichlorophenol-d3                 (42-98)
         SDMC8  (4CA)   '=  4-Chloroaniline-d4                    (8-70)

         #  Column to be used to flag recovery  values
         *  Values outside of contract required QC limits
         D  DMC diluted out
         Page 	 of 	
                                      FORM II LCSV-1
                                                                                  OLC03.2

-------
                                           2 LCD
       LOW CONCENTRATION WATER SEMIVOLATILE DEUTERATED MONITORING COMPOUND RECOVERY •
Lab Name:

Lab Code:
Contract:
                       Case  No.:
Client No.:
                                                                SDG No.:

)1
)2
)3
14
)5
)6
)7
)8
19
LO
LI
L2
L3
L4
L5
L6
L7
L8
L9
>0
!1
22
23
24
25
>e
27
28
29
30
EPA
SAMPLE NO.






























SDMC9
(DMP) #






























SDMC10
(ACY) #






























SDMC11
(4NP) #






























SDMC12
(FLR) #






























SDMC13
(NMP) #






























SDMC14
(ANC)#






























SDMC15
(PYR) #






























SDMC16
(BAP) #






























TOT
OUT






























        SDMC9  (DMP)   = Dimethylphthalate-d6
        SDMCIO(ACY)   = Acenaphthylene-d8
        SDMC1K4NP)   = 4-Nitrophenol-d4
        SDMC12(FLR)   = Fluorene-dlO
        SDMC13(NMP)   = 4,6-Dinitro-methylphenol-d2
        SDMC14(ANC)   = Anthracene-dlO
        SDMC15(PYR)   = Pyrene-dlO
        SDMC16(BAP)   = Benzo(a)pyrene-dl2
                  PC LIMITS
                  (62-102)
                  (49-98)
                  (9-181)
                  (50-97)
                  (53-153)
                  (55-116)
                  (47-114)
                  (54-120)
           Column to be used to flag recovery values .
           Values outside of contract required QC limits
           DMC diluted out
        Page 	 of 	
                                       FORM II  LCSV-2
                                                                                  OLC03.2

-------
                               2LCE
       LOW CONCENTRATION WATER PESTICIDE SURROGATE RECOVERY
Lab Name :
Lab Code :
GC Column (1) :
Case No. :
ID: IVtt-
                                 Contract :
                                 Client  No. :
                           (MM)   GC  Column (2) :
      SDG No.:

     	 ID:
                                                                      (MM)

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























TCX 1
%REC #






























TCX 2
%REC #






























DCB 1
%REC #






























DCB 2
%REC #






























OTHER
(1)






























OTHER
(2)






























TOT
OUT






























TCX    =   Tetrachloro-m-xylene
DCB    =   Decachlorobiphenyl

#  Column  to be used to flag recovery values
*  Values  outside of QC limits
D  Surrogate diluted out
PC I.IMITS
(30-150)
(30-150)
Page 	 of 	
                           FORM  II LCP
                                                                      OLC03.2

-------
                                           3LCA
                             LOW CONCENTRATION WATER VOLATILE
                       MATRIX SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
Lab Name:

Lab Code:
                      Case No.:
  Contract:

Client No.:
                                                               SDG No.
Matrix Spike - EPA Sample No.:
COMPOUND
1, 1-Dichloroethene
.Benzene
Trichloroethene
Toluene
Chlorobenzene
SPIKE
ADDED
(UG/L)





SAMPLE
CONCENTRATION
(UG/L)





MS
CONCENTRATION
(UG/L)





MS
%
REC #





QC
LIMITS
REC.
61-145
76-127
71-120
76-125
75-130
COMPOUND
1 , 1 -Dichloroethene
Benzene
Trichloroethene
Toluene
Chlorobenzene
SPIKE
ADDED
(UG/L)





MSD
CONCENTRATION
(UG/L)





MSD
% .
REC #





RPD #





QC
RPD
*14 .
11
14
13
13
LIMITS
REC.
61-145
76-127
71-120
76-125
75-130
 #  Column to be used to flag recovery and RPD values with an asterisk

 *  Values outside of QC limits

 RPD:  	 out of 	 outside limits
 Spike Recovery: 	 out of 	 outside limits
 COMMENTS:
                                        FORM  III LCV
                                                                                  OLC03.2

-------
                                           3LCB
                           LOW CONCENTRATION WATER  SEMIVOLATILE
                       MATRIX SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
Lab Name:
Lab Code:
                      Case No.:
Matrix Spike - EPA Sample No.:
                                             Contract:
                  Client  No.:
                                                                 SDG No.
COMPOUND
Phenol
2 - Chlorophenol
N-Nitroso-di-n-prop. (1)
4 -Chloro-3 -methylphenol
Acenaphthene
4 -Nitrophenol
2 , 4 -Dinitrotoluene
Pentachlorophenol
Pyrene
SPIKE
ADDED
(UG/L)









SAMPLE
CONCENTRATION
(UG/L)





/



MS
CONCENTRATION
(UG/L)









MS
%
REC #









QC
LIMITS
REC.
12-110
27-123
41-116
23-97
46-118
10-80
24-96
9-103
26-127
COMPOUND
Phenol
2 -Chlorophenol
N-Nitroso-di-n-prop. (1)
4 - Chloro-3 -methylphenol
Acenaphthene
4 -Nitrophenol
2 , 4 -Dinitrotoluene
Pentachlorophenol
Pyrene
SPIKE
ADDED
(UG/L)









MSD
CONCENTRATION
.(UG/L)









MSD
%
REC #









RPD #









QC L
RPD
42
40
38
42
31
50
38
50
31
IMITS
REC.
12-110
27-123
41-116
23-97
46-118
10-80
24-96
9-103
26-127
 (1) N-Nitroso-di-n-propylamine
# Column  to be used  to  flag  recovery and RPD values with an asterisk

* Values  outside of  QC  limits
            out  of
RPD: 	
Spike Recovery:

COMMENTS:   	
	 outside limits
 out of 	 outside  limits
                                       FORM III LCSV
                                                                                 OLC03.1

-------
                                          3LCC
                           LOW CONCENTRATION WATER PESTICIDE
                      MATRIX  SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
Lab Name:
Lab Code:
Matrix Spike -
Instrument ID :
Contract :
Case No.: Client No.: SDG No.:
EPA Sample No. :
GC Column: ID:


(mm)
COMPOUND
gamma-BHC
(Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4,4 ' -DDT
SPIKE
ADDED .
(UG/L)






SAMPLE
CONCENTRATION
(UG/L)






MS
CONCENTRATION
(UG/L)






MS
%
REC #






QC
LIMITS
REC.
56-123
40-131
40-120
52-126
56-121
38-127 .
COMPOUND
gamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4,4' -DDT
SPIKE
ADDED
(UG/L)






MSD
CONCENTRATION
(UG/L)






MSD
%
REC #






RPD #






QC
RPD
15
20
22
18
21
27
LIMITS
REC.
56-123
40-131
40-120
52-126
56-121
38-127
# Column to be used to flag recovery and RPD values with an asterisk

* Values outside of QC limits

RPD: 	 out of 	 outside limits
Spike Recovery:	 out of 	 outside limits
COMMENTS:
Page 	 of 	
                                      FORM III LCP-1
                                                                                 OLC03.2

-------
                                           3 LCD
                       LOW CONCENTRATION WATER PESTICIDE LAB CONTROL
                                     SAMPLE RECOVERY
                                                                             EPA SAMPLE NO.
Lab Name:
Lab Code:
                                                Contract:
                      Case No.:
                Client  No.
                                                               SDG No.:
Lab Sample ID:
LCS Aliquot: _
                                                       LCS Lot No.:
.(UL)
Concentrated Extract Volume: 	
Injection Volume: 	(UL)
            .(UL)
Date Extracted:
Date Analyzed:
Sulfur Cleanup:  (Y/N)
Instrument ID  (1): 	
                            Dilution  Factor:
    GC  Column  (1):
   ID:
                .(MM)
COMPOUND
gamma- BHC (Lindane)
Heptachlor epoxide
Dieldrin
4,4' -DDE
Endrin
Endosulfan sulfate
gamma - Chi ordane
AMOUNT ADDED
(NG)







AMOUNT RECOVERED
(NG)







%REC #







QC LIMITS
50-120
50-150
.30-130
50-150
50-120
50-120
30-130
Instrument ID  (2):
        GC  Column (2):
                                                                ID:
                  (MM)
COMPOUND
gamma -BHC (Lindane)
Heptachlor epoxide
Dieldrin
4,4' -DDE
Endrin
Endosulfan sulfate
gamma - Chi ordane
AMOUNT ADDED
(NG)







AMOUNT RECOVERED
(NG)







%REC #







QC LIMITS
50-120
50-150
30-130
50-150
50-120
50-120
30-130
# Column to be used  to  flag recovery values  with an asterisk
* Values outside of  QC  limits
LCS Recovery: _
                      outside limits  out  of
                                                   total.
                                      FORM III LCP-2
                                                                                  OLC03.2

-------
Lab Name:
Lab Code:
Lab Sample ID:
Lab File ID:
Instrument ID:
GC Column: 	
 COMMENTS:
 Page 	 of 	
                                           4LCA
                                 LOW CONCENTRATION WATER
                                   VOLATILE METHOD BLANK SUMMARY
                                              EPA SAMPLE NO.
                                         Contract:
                       Case No.:
                                           Client No.:
                                                                SDG No.:
                       Date Analyzed:

                       Time Analyzed:
ID:
(MM)
Length:
.(M)
                THIS METHOD BLANK APPLIES TO THE FOLLOWING SAMPLE ANALYSES:-

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























LAB
SAMPLE ID






























'LAB
FILE ID






























TIME
ANALYZED






























                                        FORM IV LCV
                                                                                  OLC03.1

-------
Lab Name:
Lab Code:
                                           4LCB
                                 LOW CONCENTRATION WATER
                                 SEMIVOLATILE METHOD BLANK SUMMARY
                                                                           EPA SAMPLE NO.
                                         "Contract:
          	  Case No. :  	  Client No. :  	  SDG No. :

Lab Sample ID: 	                        Date  Extracted: 	

Lab File ID: 	                             Date  Analyzed: 	

Instrument ID: 	                           Time  Analyzed: 	
                   THIS METHOD BLANK APPLIES TO THE FOLLOWING ANALYSES:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























LAB
SAMPLE ID






























LAB
FILE ID






























DATE
ANALYZED






























COMMENTS:
Page 	 of 	
                                       FORM IV LCSV
                                                                                 OLC03.2

-------
                                          4LCC
                                 LOW CONCENTRATION WATER
                                    PESTICIDE METHOD BLANK SUMMARY
                                                                               EPA SAMPLE NO.
Lab Name:
Lab Code:
                                          Contract:
                       Case No.:
                   Client  No.:
                                         SDG No.:
Date Extracted:
                             Lab Sample ID:
Date Analyzed  (1) :
Time Analyzed  (1) :
Instrument ID  (1):
GC Column  (1): 	
ID:
Sulfur Cleanup:  (Y/N)
                    Date Analyzed (2):  	
                    Time Analyzed (2):  	
                    Instrument ID (2):  	
.(MM)  GC Column  (2) : 	  ID:  	(MM)
                    Extraction (Sepf/Cont):  	
               THIS METHOD BLANK APPLIES TO THE FOLLOWING SAMPLE ANALYSES:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
EPA
SAMPLE NO.


























LAB
SAMPLE ID


























DATE
ANALYZED 1


























DATE
ANALYZED 2


























 COMMENTS:
 Page 	 of 	
                                        FORM IV LCP
                                                                                  OLC03.2

-------
                                           5LCA
          LOW CONCENTRATION WATER VOLATILE ORGANIC INSTRUMENT PERFORMANCE CHECK
                                 BROMOFLUOROBENZENE  (BFB)
Lab Name:
Lab_Code:.
                                             Contract:
                     _Case_No.._:__
                    Client_No_._:_
Lab File ID:
Instrument ID:
GC Column: 	.
ID:
            .(MM)
BFB Injection Date:
BFB Injection Time:
Column Length: 	
m/e
50
75
95
96
173
174
175
176
177
ION ABUNDANCE CRITERIA
8.0 - 40.0% of mass 95
30.0 - 66.0% of mass 95
Base peak, 100% relative abundance
5.0 - 9.0% of mass 95
Less than 2.0% of mass 174
50.0 - 120.0% of mass 95
4.0 - 9.0 % of mass 174
93.0 - 101.0% of mass 174
5.0 - 9.0% of mass 176
% RELATIVE
ABUNDANCE




( )1

( )1
( )1
( )2
         1-Value is  % mass  174
                                  2-Value is % mass  176
           THIS  CHECK APPLIES TO THE FOLLOWING SAMPLES, BLANKS, AND STANDARDS:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
EPA
SAMPLE NO.






















LAB
SAMPLE ID






















LAB
FILE ID






















DATE
ANALYZED






















TIME
ANALYZED











•










Page 	 of 	
                                        FORM V LCV
                                                                                  OLC03.2

-------
                                           5LCB
        LOW CONCENTRATION WATER SEMIVOLATILE ORGANIC INSTRUMENT PERFORMANCE CHECK
                           DECAFLUOROTRIPHENYLPHOSPHINE (DFTPP)
Lab Name:

Lab Code:
Contract:
                      Case No.:
                                             Client No.:
                                                                     SDG No.:
Lab File ID:
Instrument ID:
DFTPP Injection Date:

DFTPP Inj ection Time:
m/e
51
68
69
70
127
197
198
199
275
365
441
442
443
ION ABUNDANCE CRITERIA
30.0- 80.0% of mass 198
Less than 2.0% of mass 69
Mass 69 relative abundance
Less than 2.0% of mass 69
25.0 - 75.0% of mass 198
Less than 1.0% of mass 198
Base Peak, 100% relative abundance
5.0 to 9.0% of mass 198
10.0- 30.0% of mass 198
Greater than 0.75% of mass 198
Present, but less than mass 443
40.0 - 110.0% of mass 198
15.0 - 24.0% of mass 442
% RELATIVE ABUNDANCE

( )1

( )1








( )2
         1-Value is % mass 69
            2-Value is % mass 442
            THIS CHECK APPLIES TO THE FOLLOWING SAMPLES,  BLANKS,  AND STANDARDS:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
EPA
SAMPLE NO.






"•















LAB
SAMPLE ID






















LAB
FILE ID






















DATE
ANALYZED






















TIME
ANALYZED






















 Page 	 of 	
                                        FORM V LCSV
                                                                                  OLC03.2

-------
                                           6LCA
            LOW CONCENTRATION WATER VOLATILE ORGANICS INITIAL CALIBRATION DATA
Lab Name:

Lab Code:
Instrument ID:
GC Column:
                      Case No.:
                         ID:
 Calibration Date(s):

 Calibration Times:

	(MM)    Length:
                                                          Contract:
             Client No.:
                                                                     SDG No.
(M)
LAB FILE ID: RRFO . 5
RRF5 = RRF

COMPOUND
Dichlorodifluoromethane
Chloromethane
Vinyl Chloride *
Bromomethane *
Chloroethane
Trichlorof luoromethane
1, 1-Dichloroethene *
1, 1,2-Trichloro-
1,2 , 2-trif luoroethane
Acetone
Carbon Disulfide
Methyl Acetate
Methylene Chloride
trans -1,2 -Dichloroethene
Methyl tert -Butyl Ether
1 , 1 -Dichloroethane *
cis- 1,2 -Dichloroethene
2-Butanone
Bromochloromethane *
Chloroform *
1,1,1-Trichloroethane *
Cyclohexane
Carbon Tetrachloride *
Benzene *
1 , 2 -Dichloroethane *
Trichloroethene *
Me t hy 1 eye 1 ohexane
10
RRFO. 5




























RRF1



























RRF5


























RRF1
RRF25
RRF10


























=
=

RRF25



























RRF


























%
RSD


*
*


*







*


*
*
*

*
*
*
*

*Compounds with  required minimum RRF  and maximum %RSD values.
All other compounds must meet a minimum RRF  of  0.010.
                                       FORM VI  LCV-1
                                                                                 OLC03.2

-------
                                          6LCB
           LOW CONCENTRATION WATER VOLATILE ORGANICS INITIAL CALIBRATION DATA
Lab Name :
Lab Code:
Instrument ID :
GC Column:
Contract :
Case No. : Client No. : _...
Calibration Date(s) :
Calibration Times:
ID: (MM) Lenath:

SDG No. :


(M)
LAB FILE ID: RRF0.5
RRF5 - RRF10
COMPOUND
1 , 2 -Dichloropropane
Bromodichloromethane *
cis-l,3-Dichloropropene *
4 -Methyl - 2 -pentanone
Toluene *
trans-l,3-Dichloropropene *
1, 1, 2-Trichloroethane *
Tetrachloroethene *
2-Hexanone
Dibromochloromethane *
1 , 2 -Dibromoethane *
Chlorobenzene *
Ethylbenzene - *
Xylene (total) *
Styrene *
Bromoform *
I sopropylbenzene
1,1,2,2 -Tetrachloroethane *
1 , 3 -Dichlorobenzene *
1 , 4 -Dichlorobenzene *
1 , 2 -Dichlorobenzene *
l,2-Dibromo-3-chloropropane
1,2,4 -Tri chlorobenzene *
1,2,3 -Trichlorobenzene *
RRF0.5


























RRF1
























RRF5




j



















RRF1
RRF25 =
RRF10
























=

RRF25

























RRF
























%
RSD

*
*

*
*
*
*

*
*
*
*
*
*
*

*
*
*
*

*
*
*Compounds with required minimum RRF and maximum %RSD values.
All other compounds must meet a minimum RRF of 0.010.
                                      FORM VI LCV-2
                                                                                 OLC03.2

-------
                                           6LCC
            LOW CONCENTRATION WATER VOLATILE ORGANICS INITIAL CALIBRATION DATA
Lab Name:

Lab Code:
Instrument ID:
GC Column:
                                             Contract:
                      Case No.:
             Client No.:
                         ID:
 Calibration Date(s):

 Calibration Times:

	(MM)   Length:
                                                                      SDG NO.:
.(M)
LAB FILE ID: RRF0.5 = RRF1
RRF5 - ' RRF10 - RRF25 -
COMPOUND
Vinyl chloride-d3
Chloroethane-d5
1, l-Dichloroethene-d2
2 - Butanone - d5
Chloroform-d
1, 2-Dichloroethane-d4
Benzene-d6
1 , 2 -Dichloropropane-d6
Jpluene-d8
.. trans -1,3-Dichl oropropene - d4
, 2 _- Hexanone - d5
,Bromoform-d
, i, 1 , 2 , 2-Tetrachloroethane-d2
1, 2-Dichlorobenzene-d4
RRF0.5














RRF1














RRF5














RRF10














RRF25





•








RRF














%
RSD














*Compounds with  required minimum RRF  and maximum %RSD values.
All other compounds must meet a minimum RRF of  0.010.
                                       FORM VI LCV-3
                                                                                  OLC03.2

-------
                                           6 LCD
          LOW CONCENTRATION WATER SEMIVOLATILE ORGANICS INITIAL CALIBRATION DATA
Lab Name:

Lab Code:
                                             Contract:
                      Case No.:
               Client No.:
                                                                      SDG No.:
Instrument ID:
Calibration Date(s):

Calibration Times:
LAB FILE ID: RRF5
RRF20 = RRF50

COMPOUND
Benzaldehyde
Phenol *
bis- (2-Chloroethyl) ether *
2-Chlorophenol *
2-Methylphenol *
2,2' -oxybis (1-Chloropropane)
Acetophenone
4 -Methylphenol *
N-Nitroso-di-n-propylamine • *
Hexachloroethane *
Nitrobenzene *
Isophorone *
2-Nitrophenol *
2 , 4 -Dimethylphenol *
bis (2-Chloroethoxy)methane *
2,4-Dichlorophenol *
Naphthalene *
4-Chloroaniline
Hexachlorobutadiene
Caprolactam
4-Chloro-3-methylphenol *
2-Methylnaphthalene *
Hexachlorocyclopentadiene
2,4,6 -Trichlorophenol *
2,4,5-Trichlorophenol *
1,1' -Biphenyl
2-Chloronaphthalene *
2-Nitroaniline
Dimethylphthalate
2 , 6 -Dinitrotoluene *
Acenaphthylene *
3 -Nitroaniline
Acenaphthene *
2 , 4 -Dinitrophenol
4 -Nitrophenol
Dibenzofuran *

_

RRF5







































RRF10







































RRF20




































RRF10
RRF80

RRF50





































_

RRF80







































RRF






































%
RSD

*
*
*
*


*
*
*
it
if
*
*
• *
. *
*



*
*

*
+

*


*
*

*


*
 * Compounds with required minimum RRF and maximum %RSD values.
 All other compounds must meet a minimum RRF of 0.010.
                                       FORM VI  LCSV-2
                                                                                  OLC03.2

-------
                                           6LCE
          LOW CONCENTRATION WATER SEMIVOLATILE ORGANICS INITIAL CALIBRATION DATA
Lab Name:

Lab Code:
                                             Contract:
                      Case No.:
               Client No.:
                                                                      SDG No.:
Instrument ID:
Calibration Date(s):

Calibration Times:
LAB FILE ID: RRF5
RRF20 = RRF50

COMPOUND
2,4-Dinitrotoluene *
Diethylphthalate
Fluorene *
4-Chlorophenyl-phenylether *
4-Nitroaniline
4, 6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine (1)
1,2,4,5 Tetrachlorobenzene
4-Bromophenyl-phenylether *
Hexachlorobenzene *
Atrazine
Pentachlorophenol *
Phenanthrene . *
Anthracene *
Di-n^butylphthalate
Fluoranthene *
Pyrene *
Butylbenzylphthalate
3,3' -fiichlorobenzidine
Benzo (a) anthracene *
Chrysene *
bis (2-Ethylhexyl)phthalate
Di-n-octylphthalate
Benzo (b) fluoranthene *
Benzo ( k ) f luoranthene *
Benzo ( a ) pyrene *
Indenod, 2, 3-cd)pyrene *
Dibenzo (a, h) anthracene *
Benzo (g, h, i)perylene *
_
=

RRF5
































RRF10
































RRF20





























RRF10
RRF80

RRF50





























__
=

RRF80
































RRF































%
RSD
*

*
*




*
*

*
*
*

*
*


*
*


*
*
*
*
*
*
 (1) Cannot be separated from Diphenylamine
* Compounds with required minimum RRF and maximum %RSD values.
All other compounds must meet a minimum RRF of 0.010.
                                      FORM VI LCSV-2
                                                                                 OLC03.2

-------
                                           6LCF
          LOW  CONCENTRATION WATER SEMIVOLATILE  ORGANICS  INITIAL  CALIBRATION DATA
Lab Name:
Lab Code:
Instrument ID:
                                             Contract:
                      Case  No.:
               Client No.:
                                        SDG No.:
Calibration Date(s):
Calibration Times:
LAB FILE ID: R
RRF20 = R
COMPOUND
Phenol -d5
bis- (2-Chloroethyl) ether-
d8
2-Chlorophenol-d4
4 -Methylphenol - d8
Nitrobenzene-d5
2-Nitrophenol-d4
2,4-Dichlorophenol-d3
"4-Chloroaniline-d4
Dimethylphthalate-d6
Acenaphthylene - d8
4 - Ni t ropheno 1 - d4
Fluorene-dlO
4 , 6-Dinitro-methylphenol-
d2
Anthracene-dlO
Pyrene-dlO
Benzo (a) pyrene-d!2
RF5
RF50
RRF5
















X
=
RRF10


















RRF20
















RRF10
RRF80
RRF50
















_
-
RRF80


















RRF


















%
RSD
















                                       FORM VI  LCSV-3
                                                                                  OLC03.2

-------
                                           6LCG
                            LOW CONCENTRATION WATER PESTICIDE
                     INITIAL CALIBRATION OF SINGLE COMPONENT ANALYTES
Lab Name:

Lab Code:
Instrument ID:

GC Column: 	
                      Case No.:
                                             Contract:
                Client No.:
                                                           mid
Level (x low):  low 	

ID: 	(MM)   Date(s) Analyzed:
 SDG No.:

	 high
COMPOUND
alpha-BHC
beta-BHC
delta-BHC
gamma -BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4,4' -DDE
Endrin '
Endosulfan II
4,4' -DDD
Endosulfan sulfate
4,4' -DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha - Chi ordane
gamma - Chi ordane
Tetrachloro-m-xylene
Decachlorobiphenyl
RT
LOW






















OF STANDAF
MID






















JDS
HIGH






















MEAN
RT






















RT W
FROM






















ENDOW
TO






















* Surrogate retention times are measured from Standard Mix A analyses.

Retention time windows are ±0.05 minutes for all compounds that elute before
Heptachlor expoxide, ±0.07 minutes for all other compounds, except ±0.10
minutes for Decachlorobiphenyl.
                                      FORM VI LCP-1
                                                                                 OLC03.2

-------
                                           6LCH
                            LOW CONCENTRATION WATER PESTICIDE
                     INITIAL CALIBRATION OF SINGLE COMPONENT ANALYTES
Lab Name:

Lab Code:
Instrument ID:

GC Column: 	
                                             Contract:
                      Case No.:
                Client No.:
                                                           mid
Level (x low):  low    '	

ID: 	(MM)   Date(s)  Analyzed:
 SDG NO.:

	 high
COMPOUND
alpha-BHC
beta-BHC
delta-BHC
gamma -BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4,4' -DDE
Endrin
Endosulfan II
4,4' -DDD
Endosulfan sulfate
4,4' -DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha-Chlordane •
gamma - Chi or dane
Tetrachloro-m-xylene
Dec achlorobiphenyl
LOW






















CALIBRATIC
MID






















)N FACTORS
HIGH






















MEAN






















%RSD






















 *  Surrogate calibration factors are measured from Standard Mix A analyses.
                                       FORM VI LCP-2
                                                                                  OLC03.2

-------
                                           6LCI
                            LOW CONCENTRATION WATER PESTICIDE
                      INITIAL CALIBRATION OF MULTICOMPONENT ANALYTES
Lab Name:

Lab Code:
                                             Contract:
                      Case No.:
               Client No.:
                                                                     SDG No.:
Instrument ID:

GC Column: 	
Date(s)  Analyzed:

ID: 	(MM)
COMPOUND
Toxaphene
Aroclor 1016
Aroclor 1221
'.'Aroclor 1232
•f. f
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
AMOUNT
(NG)

















PEAK1
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
RT

















i






















RT WI
FROM








































NDOW
TO








































CALIBRATION
FACTOR











N




























*At least 3 peaks for each column are required for identification of
multicomponent analytes.
                                       FORM VI  LCP-3
                                                                                 OLC03.2

-------
                                           6LCJ
               LOW CONCENTRATION WATER PESTICIDE ANALYTE RESOLUTION SUMMARY
Lab Name:
Lab Code:
                      Case No.
                                             Contract:
               Client No.:
                                                                      SDG No.
GC Column  (1) .-
ID:
EPA Sample No.  (RESC##):

Date Analyzed  (1):    .
.(MM)
Instrument ID (1):

Lab Sample ID (1):

Time Analyzed (1):

01
02
03
04
05
06
07
08
09
ANALYTE









RT









RESOLUTION









mn (2): ID: (MM) Instrument ID (2):
pie No. (RESC«#) : Lab Samole ID (2) :
alyze

01
02
03
04
05
06
07
08
09
sd (2) : -Time Analyzed (2) :

ANALYTE









RT









RESOLUTION
(%)


\,







                                        FORM VI  LCP-4
                                                                                   OLC03.2

-------
                            6LCK
LOW CONCENTRATION WATER PERFORMANCE EVALUATION MIXTURE  (PEM)
Lab >
Lab C
GC Cc
EPA £
Date
lame:
:ode:
Contract :
Case No. : Client No. : SDG No. :

jlumn (1) : ID: (MM) Instrument ID (1) :
3amp]
Anal

01
02
03
04
05
06
07
08
e No. (PEMttft) : Lab Santole ID (1) :
vzed (1) : Time Analyzed (1) :

ANALYTE








RT








RESOLUTION
(%)









Dlumr
Sampl
Anal
•'

01
02
03
04
05
06
07
08
L (2) : ID: (MM) Instri
e No. (PEM##) : Lab S<
yzed (2) : Time ;

ANALYTE








ament ID (2) :
ample ID (2) :
Analyzed (2) :

'RT












RESOLUTION
(%)





















                        FORM VI LCP-5
                                                                   OLC03.2

-------
                                           6LCL
               LOW CONCENTRATION WATER INDIVIDUAL STANDARD MIXTURE A  (INDA)
Lab Name:

Lab Code:
                      Case No.:
GC Column  (1) :
ID:
EPA Sample No.  (INDAM##):

Date Analyzed  (1): 	
               Contract:
               Client No.:
.(MM)
Instrument ID  (1):

Lab Sample ID  (1):

Time Analyzed  (1):
                                                                      SDG No.

01
02
03
04
05
06
07
08
09
10
11
ANALYTE










•
RT











RESOLUTION











Dlumn
3ampl
Anal


01
02
03
04
05
06
07
08
09
10
11
(2) : ID: (MM) Instn.
e NO. (INDAM##) : ' Lab Sc
vzed (2) : Time 1

ANALYTE











iment ID (2) :
imple ID (2) :
taalyzed (2) :

RT















RESOLUTION
(%)



























                                       FORM VI LCP-6
                                                                                  OLC03.2

-------
                                           6LCM
               LOW CONCENTRATION WATER INDIVIDUAL STANDARD  MIXTURE B (INDB)
Lab Name:

Lab Code:
                                             Contract:
                      Case No.:
       Client No.:
GC Column  (1):
                              ID:
.(MM)
EPA Sample No.  (INDBM##):

Date Analyzed  (1): 	
Instrument ID  (1):
Lab Sample ID  (1):

Time Analyzed  (1):
                                                                      SDG No.

01
02
03
04
05
06
07
08
09
•10
-11
-12
,..13
ANALYTE













RT













RESOLUTION



•









GC Column  (2) :
                              ID:
EPA Sample No.  (INDBMtti):

Date .Analyzed. (2) : 	
.(MM)
Instrument ID  (2):

Lab Sample ID  (2):

Time Analyzed  (2):

01
02
03
04
05
06
07
08
09
10
11
12
13
ANALYTE













RT













RESOLUTION













                                       FORM VI LCP-7
                                                                                  OLC03.2

-------
                                           7LCA
              LOW CONCENTRATION WATER VOLATILE CONTINUING CALIBRATION CHECK
Lab Name:

Lab Code:
                                             Contract:
                      Case No.:
           Client No.:
                                                                     SDG No.
Instrument ID:

Lab File ID:
                            Calibration Date:
                            Time:
EPA Sample No.(VSTD005##):

GC Column: 	    ID:
.(MM)
    Init.  Calib.  Date(s):

    Init.  Calib.  Times:

Length: 	 (M)
COMPOUND
Dichlorodifluoromethane
Chloromethane
Vinyl Chloride
Bromomethane
Chloroethane
Trichlorofluoromethane
1 , 1 -Dichloroethene
l,l,2-Trichloro-l,2,2-trifluoroethane
Acetone
Carbon Disulfide
Methyl Acetate
Methylene Chloride
trans-l, 2 -Dichloroethene
Methyl tert -Butyl Ether
1, 1-Dichloroethane
cis -1,2 -Dichloroethene
2-Butanone
Bromochloromethane
Chloroform
1,1, 1-Trichloroethane
Cyclohexane
Carbon Tetrachloride
Benzene
1 , 2 -Dichloroethane
Trichlor.oethene
Methyl cyclohexane
RRF


























RRF5


























MIN
RRF


0.100
0.100


0.100







0.200


0.050
0.200
0.100

0.100
0.400
0.100
0.300

%D


























MAX
%D


30.0
30.0


30.0







30.0


30.0
30.0
30.0

30.0
30.0
30.0
30.0

 All other compounds must meet a minimum RRF of 0.010.
                                       FORM VII  LCV-1
                                                                                  OLC03.2

-------
                                           7LCB
              LOW CONCENTRATION WATER VOLATILE CONTINUING CALIBRATION CHECK
Lab Name:
Lab Code:
Instrument ID:
Lab File ID:
                                             Contract:
                      Case No.:
             Client No.:
                                                                     SDG No.
EPA Sample No.(VSTD005##):

GC Column:	ID:.
                               Calibration Date:
                                                               Time:
.(MM)
Init. Calib. Date(s):

Init. Calib. Times:

Length:
(M)
COMPOUND
1 , 2 -Dichloropropane
Bromodichloromethane
cis-1, 3-Dichloropropene
4 -Methyl - 2 -pentanone
Toluene
trans -1 , 3 -Dichloropropene
1,1, 2-Trichloroethane
Tetrachloroethene
2-Hexanone
Dibromochloromethane
1 , 2 -Dibromoethane
•H i-i -
Chlqrobenzene
Ethylbenzene
Xylene (total)
Styrene
Bromof orm
I sopropylbenzene
1,1,2,2 -Tetrachloroethane
1 , 3 -Dichlorobenzene
1 , 4 -Dichlorobenzene
1 , 2 -Dichlorobenzene
1, 2-Dibromo-3-chloropropane
1,2 , 4-Trichlorobenzene
1,2,3 -Trichlorobenzene
RRF '
























RRF5
























MIN
RRF

0.200
0.200

0.400
0.100
0.100
0.100

0.100
0.100
0.500
0.100
0.300
0.300
0.050

0.100
0.400
0.400
0.400

0.200
0.200
%D
























MAX -
%D

30.0
30.0

30.0
30.0
30.0
30.0

30.0
30.0
30.0
30.0
30.0
30.0
30.0

30.0
30.0
30.0
30.0

30.0
30.0
All other compounds must  meet  a  minimum RRF of  0.010.
                                      FORM VII LCV-2
                                                                                 OLC03.2

-------
                                           7LCC
              LOW CONCENTRATION WATER VOLATILE .CONTINUING  CALIBRATION  CHECK
Lab Name:

Lab Code:
Instrument ID:

Lab File ID:
                                             Contract:
                      Case No.:
             Client No.:
                                                                      SDG No.
EPA Sample No.(VSTD005##):

GC Column:	ID:
                               Calibration Date:
                                                               Time:
.(MM)
Init. Calib. Date(s): 	

Init. Calib. Times:	

Length: 	(M)
COMPOUND
Vinyl Chloride-d3
Chloroethane-d5
1 , l-Dichloroethene-d2
2-Butanone-d5
Chloroform-d
1 ; 2-Dichloroethane-d4
Benzene -d6
1 , 2-Dichloropropane-d6
Toluene -d8
trans- 1, 3-Dichloropropene-d4
2-Hexanone-d5 >•
Bromoform-d
1, l,2,2-Tetrachloroethane-d2
1 , 2 -Dichlorobenzene-d4
RRF



-
-









RRF5














MIN
RRF














%D














MAX
%D














                                       FORM VII LCV-3
                                                                                  OLC03.2

-------
                                           7LCD
            LOW CONCENTRATION WATER  SEMIVOLATILE  CONTINUING CALIBRATION CHECK
Lab Name:
Lab Code:
                                             Contract:
                      Case No.:
       Client No.:
                                                                      SDG No.:
Instrument ID:
Lab File ID:
                            Calibration Date:
                     Time:
EPA Sample No.(SSTD020##):
GC Column: 	 ID: 	
       Init. Calib. Date(s):.
       Init. Calib. Times:
.(MM)
COMPOUND
Benzaldehyde
Phenol
bis- (2-Chloroethyl)ether
2 - Chlorophenol
2 -Methylphenol
2,2' -oxybis (1-Chloropropane)
Acetophenone
4 -Methylphenol
N-Nitroso-di-n-propylamine
Hexachloroethane
Nitrobenzene
Isophorone
2 -NiCrophenol
2 , 4 -Dimethylphenol
bis (2-Chloroethoxy) methane
2 , 4 -Dichlorophenol
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
Caprolactam
4 - Chloro- 3 -methylphenol
2 -Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6-Trichlorophenol
2 , 4 , 5-Trichlorophenol
1,1' -Biphenyl •
2-Chloronaphthalene
2-Nitroaniline
Dimethylphthalate
2 , 6 -Dinitrotoluene
Acenaphthylene
3 -Nitroaniline
Acenaphthene
2 , 4 -Dinitrophenol
4 -Nitrophenol
Dibenzofuran
RRF




































RRF20




































MIN
RRF

0.800
0.700
0.800
0.700


0.600
0.500
0.300
0.200
0.400
0.100
0.200
0.300
0.200
0.700



0.200
0.400

0.200
0.200

0.800


0.200
0.900

0.900


0.800
%D




































MAX
%D

25.0
25.0
25.0
25.0


25.0
25.0
25.0
25.0
25.0
30.0
30.0
25.0
25.0
25.0



25.0
25.0

25.0
25.0

25.0


25.0
25.0

25.0


25.0
All other compounds  must  meet  a minimum RRF  of  0.010.
                                      FORM VII  LCSV-1
                                                                                  OLC03.2

-------
                                           7LCE
            LOW CONCENTRATION WATER SEMIVOLATILE CONTINUING CALIBRATION CHECK
Lab Name:

Lab Code:
    Contract:
                      Case No.:
    Client  No.:
                                                                     SDG No.:
Instrument ID:

Lab File ID:
                              Calibration Date:
                      Time:
EPA Sample No,(SSTD020##): 	

GC Column: 	 ID:
    Init.  Calib.  Date(s):

    Init.  Calib.  Times:
.(MM)
COMPOUND
2 , 4 -Dinitrotoluene
Diethylphthalate
Fluorene
4 - Chlorophenyl -phenylether
4-Nitroaniline
4 , 6 -Dinitro-2 -methylphenol
N-Nitrosodiphenylamine (1)
1,2,4,5 Tetrachlorobenzene
4 -Bromophenyl -phenylether
Hexachlorobenzene
Atrazine
Pentachlorophenol
Phenanthrene
Anthracene
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Benzo (a) anthracene
Chrysene
bis (2-Ethylhexyl)phthalate
Di-n-octylphthalate
Benzo (b) fluoranthene
Benzo (k) fluoranthene
Benzo (a) pyrene
Indeno ( 1 , 2 , 3 - cd ) pyrene
Dibenzo (a, h) anthracene
Benzo (g , h , i ) perylene
RRF





























RRF20





























MIN
RRF
0.200

0.900
0.400




0.100
0.100

0.050
0.700
0.700

0.600
0.600


0.800
0.700


0.700
0.700
0.700
0.500
0.400
0.500
%D





























MAX
%D
30.0

25.0
25.0




25.0
25.0

25.0
25.0
25.0

25.0
25.0


25.0
25.0


25.0
25.0
25.0
25.0
25.0
25.0
 (1)  Cannot be separated from Diphenylamine
 All  other compounds must meet a minimum RRF of 0.010.
                                      FORM VII LCSV-2
                                                                                  OLC03.2

-------
                                           ,7LCF
            LOW CONCENTRATION WATER SEMIVOLATILE  CONTINUING  CALIBRATION CHECK
Lab Name:

Lab Code:
                                             Contract:
                      Case No.:
    Client No.:
                                                                      SDG No.:
Instrument ID:

Lab File ID:
                              Calibration Date:
                      Time:
EPA Sample No.(SSTD020##): 	

GC Column: 	 ID:
    Init.  Calib.  Date(s):

    Init.  Calib.  Times:
.(MM)
COMPOUND
Phenol -d5
bis- (2-Chloroethyl)ether-d8
2 - Chlorophenol - d4
4 -Methylphenol -d8
Nitrobenzene-d5
2-Nitrophenol-d4
2,4-Dichlorophenol-d3
4-Chloroaniline-d4
Dimethylphthalate-d6
Acenaphthyl ene - d8
4 -Nitrophenol-d4
Fluorene-dlO
4, 6-Dinitro-methylphenol-d2
Anthracene-dlO
Pyrene-dlO
Benzo (a) pyrene-d!2
RRF
















RRF20
















MIN
RRF
















%D
















MAX
%D
















                                      FORM VII LCSV-3
                                                                                 OLC03.2

-------
                                           7LCG
            LOW CONCENTRATION WATER PESTICIDE CALIBRATION VERIFICATION SUMMARY
Lab Name:

Lab Code:


GC Column:
                       Contract:
Case No.
                       Client No.:
                                                SDG No.:
                          ID:
EPA Sample No.  (PIBLK##):

Lab Sample ID  (PIBLK): 	

EPA Sample No.  (PEM##): _

Lab Sample ID  (PEM): 	
            .(MM)    Init.  Calib.  Date(s):

            	.__   Date Analyzed: 	

            	   Time Analyzed: 	

            	   Date Analyzed: 	

            	   Time Analyzed: 	
PEM
COMPOUND
alpha-BHC
beta-BHC
gamma -BHC (Lindane)
Endrin
4,4' -DDT
Methoxychlor
RT






RT W]
FROM






[NDOW
TO






CALC
AMOUNT
(NG)






NOM
AMOUNT
(NG)






%D






 4,4'-DDT %  Breakdown (1)

 Combined %  Breakdown (1)
                       Endrin  % breakdown (1) :_
                                       FORM VII LCP-1
                                                                                  OLC03.2

-------
                                           7LCH
            LOW CONCENTRATION WATER PESTICIDE CALIBRATION VERIFICATION SUMMARY
Lab Name:

Lab Code:


GC Column:
                                             Contract:
Case No.:
                       Client No.:
                                                SDG No.:
                          ID:
EPA Sample No.  (PIBLK##):

Lab Sample ID  (PIBLK): 	
EPA Sample No.  (INDAM##):

Lab Sample ID   (INDA): 	
            .(MM)    Init.  Calib.  Date(s):

            	   Date Analyzed: 	

            	   Time Analyzed: 	

            	   Date Analyzed: 	

            	   Time Analyzed: 	
INDIVIDUAL MIX A
COMPOUND
alpha-BHC
gamma-BHC (Lindane)
Heptachlor
Endosulfan I
Dieldrin
Endrin.
4,4' -ODD
4,4' -DDT
Methoxychlor
Tetrachloro-m-xylene
Decachlorobiphenyl
RT











RT W]
FROM











[NDOW
TO











CALC
AMOUNT
(NG)











NOM
AMOUNT
(NG)











%D











EPA Sample No.  (INDBM##)

Lab Sample ID   (INDB):  _
                       Date Analyzed:

                       Time Analyzed:
INDIVIDUAL MIX B
COMPOUND
beta-BHC
delta-BHC
Aldrin
Heptachlor epoxide
4,4' -DDE
Endosulfan II
Endosulfan sulfate
Endrin ketone
Endrin aldehyde
alpha - Chi ordane
gamma - Chi ordane
Tetrachloro-m-xylene
Decachlorobiphenyl
RT













RT W
FROM













INDOW
TO













CALC
AMOUNT
(NG)













NOM
AMOUNT
(NG)













%D













                                      FORM VII LCP-2
                                                                                  OLC03.2

-------
                                          8LCA
         LOW CONCENTRATION WATER VOLATILE INTERNAL STANDARD AREA AND RT SUMMARY
Lab Name:
Lab Code : Case No . :
EPA Sample No. (VSTD005##) :
Lab File ID (Standard) :
Instrument ID:
GC Column: ID:
Contract :
Client No. : SDG No. :
Date Analyzed:
Time Analyzed:

(MM) Lenqth: (M)





01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
-
12 HOUR STD
UPPER LIMIT
LOWER LIMIT
EPA SAMPLE NO.






















IS1 (CBZ)
AREA #


























RT fl


























IS2 (DFB)
AREA fl










.















RT #


























IS3 (DCB)
AREA #


























RT #


























IS1 (CBZ) = Chlorobenzene-d5
IS2 (DFB) = 1,4-Difluorobenzene
IS3 (DCB) = l,4-Dichloroberizene-d4

AREA UPPER LIMIT = +40% of internal standard area
AREA LOWER LIMIT = -40%  of internal standard area
RT UPPER LIMIT = +0.33 minutes of internal standard RT
RT LOWER LIMIT = -0.33 minutes of internal standard RT

# Column used to flag values outside QC limits with an asterisk.
* Values outside of QC limits
 Page 	of 	
                                      FORM VIII LCV
                                                                                 OLC03.2

-------
                                           8LCB
       LOW  CONCENTRATION WATER SEMIVOLATILE  INTERNAL  STANDARD AREA AND  RT  SUMMARY
Lab Name :
Lab Code : Case No . :
EPA Sample No. (SSTD020«#) :
Lab File ID (Standard) :
Instrument ID :
Contract :
Client No. :
Date Analyzed:
Time Analyzed:
GC Column:

SDG No. :


ID: (MM)

01
02
03
04
05
06
07
08
09
LO
11
12
13
14
15
16
17
18
19
20
21
22

12 HOUR STD
UPPER LIMIT
LOWER LIMIT
EPA SAMPLE NO.








""• •
• ••off
-•"
~-.r
-.
^








IS1 (DCB)
AREA #


























RT #


























IS2 (NPT)
AREA fl


























RT #


























IS3 (ANT)
AREA #


























RT #


























IS1  (DCB) = l,4-Dichlorobenzene-d4
IS2  (NPT) = Naphthalene-d8
IS3  (ANT) = Acenaphthene-dlO

AREA UPPER LIMIT = +100% of internal standard area
AREA LOWER LIMIT =  -50% of internal standard area
RT UPPER LIMIT = +0.33 minutes of internal standard RT
RT LOWER LIMIT = -0.33 minutes of internal standard RT

# Column used to flag values outside QC limits with an asterisk.
* Values outside of QC limits

Page 	 of 	
                                     FORM VIII LCSV-l
                                                                                 OLC03.2

-------
                                          8LCC
       LOW CONCENTRATION WATER SEMIVOLATILE INTERNAL STANDARD AREA' AND RT SUMMARY
Lab Name:
Lab Code : Case No . :
EPA Sample No. (SSTD020##) :
Lab File ID (Standard) :
Instrument ID:
Contract :
Client No. :
Date Analyzed:
Time Analyzed:
GC Column:

SDG No . :


ID: (MM)





01
02
03
04
05
06
07
08
09
10
11
12
L3
14
15
16
17
18
19
20
21
22

12 HOUR STD
UPPER LIMIT
LOWER LIMIT
EPA SAMPLE NO.






















IS4 (PHN)
AREA #


























RT #


























IS5 (CRY)
AREA #


























RT #


























IS6 (PRY)
AREA #


























RT #


























IS4  (PHN) = Phenanthrene-dlO
IS5  (CRY) = Chrysene-dl2
IS6  (PRY) = Perylene-dl2

AREA UPPER LIMIT = +100% of internal standard area
AREA LOWER LIMIT = -50%  of internal standard area
RT UPPER LIMIT = +0.33 minutes of internal standard RT
RT LOWER LIMIT = -0.33 minutes of internal standard RT

# Column used to flag values outside QC limits with an asterisk.
* Values outside of QC limits

Page 	 of 	
                                     FORM VIII LCSV-2
                                                                                 OLC03.2

-------
                                           8LCD
                  LOW CONCENTRATION WATER PESTICIDE ANALYTICAL SEQUENCE
Lab Name:
Lab Code :
GC Column:
Contract :
Case No. : Client No. :
ID: (MM) Init. Calib. Date(s):

SDG No . :

Instrument ID:
THE ANALYTICAL SEQUENCE OF SAMPLES, BLANKS, AND STANDARDS IS GIVEN BELOW:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
MEAN SURROGATE RT FROM INITIAL CALIBRATION
TCX : DCB :
EPA
SAMPLE NO.











.....
......



















LAB
SAMPLE ID
































DATE
ANALYZED




















v











TIME
ANALYZED

































TCX
RT #
































DCB
RT #
































TCX = Tetrachloro-m-xylene
DCB = Decachlorobiphenyl
QC LIMITS
(±0.05  MINUTES)
(±0.10  MINUTES)
# Column used to flag retention time values with an asterisk.
* Values outside of QC limits.
Page 	 of 	
                                       FORM VIII LCP
                                                                                 OLC03.2

-------
                                           9LCA
                LOW CONCENTRATION WATER PESTICIDE FLORISIL CARTRIDGE CHECK
Lab Name:

Lab Code:
                      Case No.:
Florisil Cartridge Lot Number: 	

GC Column  (1) : 	 ID:.
(MM)
                                                Contract:
          Client No.:
                                                                      SDG No.:
   Date of Analysis:

GC Column(2):  	
.(MM)
COMPOUND
alpha -BHC
qamma-BHC (Lindane)
Heptachlor
Endosulfan I
Dieldrin
Endrin
4,4' -ODD
4,4' -DDT
Methoxychlor
Tetrachloro-m-xylene
Decachlorobiphenyl
2 , 4 , 5-Trichlorophenol
SPIKE
ADDED
(NG)












SPIKE
RECOVERED
(NG)












%
REC #












QC
LIMITS
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
<5
# Column  to be  used to  flag recovery with an asterisk.
* Values  outside  of QC  limits.

                   THIS CARTRIDGE LOT APPLIES TO THE FOLLOWING SAMPLES,
                          LABORATORY CONTROL SAMPLES,  AND BLANKS:

01
02
03
• 04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
EPA
SAMPLE NO.























LAB
SAMPLE ID























DATE
ANALYZED 1























DATE
ANALYZED 2























 Page 	 of 	
                                        FORM IX LCP
                                                                                  OLC03.2

-------
Lab Name:
Lab Code:
Lab Sample ID:
Instrument ID  (1):

GC Column:(1): 	
                                           10LCA
                    LOW CONCENTRATION WATER PESTICIDE IDENTIFICATION
                                SUMMARY FOR SINGLE COMPONENT ANALYTES
                                                                            EPA SAMPLE NO.
                      Case No.:
ID:
                                             Contract:
                Client No.:
         .(MM)
Date(s) Analyzed:

Instrument ID (2) :

GC Column:(2): 	
                                                                      SDG No.:
.(MM)
ANALYTE



'_.•





COL
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
RT
















RT WI
FROM'
















NDOW
TO
















CONCENTRATION
















%D









 Page 	 of
                                       FORM X LCP-1
                                                                                  OLC03.2

-------
                                           10LCB
                    LOW CONCENTRATION WATER  PESTICIDE  IDENTIFICATION
                                 SUMMARY FOR MULTICOMPONENT ANALYTES
                                                                            EPA SAMPLE NO.
Lab Name:

Lab Code:
                Contract:
                      Case No.:
                                             Client No.:
                                                                      SDG No.:
Lab Sample ID: 	

Instrument ID  (1)

GC Column:(1): 	
ID:
	   Date(s) Analyzed:

	   Instrument ID  (2)

.(MM)   GC Column:(2): 	
                                               ID:
                                                         .(MM)
ANALYTE

COLUMN 1
COLUMN 2

COLUMN 1
COLUMN 2

COLUMN 1
COLUMN 2
PEAK
1
2
3
4
5

1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
RT



































RT WI
FROM



































NDOW
TO



































CONCENTRATION



































MEAN
CONCENTRATION









%D




 At least 3 peaks for each column are required for identification of
 multicomponent analytes.
 Page 	 of '
                                        FORM X LCP-2
                                                                                  OLC03.2

-------
                                           SAMPLE LOG-IN SHEET
Lab Name
Received By (Print Name)
Page 	 of 	 '_
Log -in Date
Received By (Signature)
Case Number
Remarks :
1. Custody Seal (s)
2 . Custody Seal
Nos.
3 . Chain of Custody
Records
4 . Traffic Reports
or Packing Lists
5. Airbill
6. Airbill No.
7 . Sample Tags
Sample Tag Nos .
8 . Sample Condition
9 . Cooler
' Temperature 	
Indicator Bottle
10. Cooler
Temperature
11. Does
information on
custody
records ,
traffic
reports,
and sample tags
agree?
12 . Date Received
at Lab
13 . Time Received

Present /Absent*
Intact /Broken

Present /Absent*
Present /Absent*
Airbill/Sticker
Present/Absent*

Present /Absent*
Listed/Not
Listed on Chain-
of- Custody
Intact /Broken* /
Leaking
Present /Absent*
Yes /No*


Sample Transfer
Fraction
Area #
By . .
On
Fraction
Area #
By
On
Sample Delivery Group No.

EPA
Sample #



















Corresponding
Sample
Tag 8



















Assigned
Lab #



















Client Number

Remarks : Condition
of Sample
Shipment , etc .






-



_^








* Contact SMO and attach record of resolution
Reviewed By
Date
Logbook No.
Logbook Page No .
                                                  FORM DC-1
                                                                                         OLC03.2

-------
                 LOW CONCENTRATION WATER ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET
LABORATORY NAME
CITY/STATE
CASE NO. SDG NO.

SDG NOS. TO FOLLOW
CLIENT NO.
CONTRACT NO.
SOW NO.
        All  documents delivered in the Complete SDG File must  be original documents where possible.
                                                                                   PAGE NOs       CHECK
                                                                                 FROM    TO     LAB    EPA
1. Inventory Sheet  (Form DC-2)  (Do Not Number)                                   	   	   	   	
2. SDG Case Narrative                                                            	   	   	   	
3. SDG Cover Sheet/Traffic Report                                                	   	   	   	
4. Volatiles Data
   a.  QC Summary
       Deuterated Monitoring Compound Recovery  (Form II LCV)                     	   	   		
       Matrix Spike/Matrix Spike Duplicate Recovery
         (Form III LCV)  (if Region requests)                                      	   	   		
       Method Blank Summary  (Form IV LCV)                                        	   	   	   	
       GC/MS Instrument Performance Check  (Form V LCV)                           	   	   	   	
       Internal Standard Area and RT Summary  (Form VIII LCV)                     	   	   	   	
    b.  Sample Data
       TCL Results -  (Form I LCV-1, LCV-2)
       Tentatively Identified Compounds  (Form I LCV-TIC)
       Reconstructed  Total Ion Chromatograms  (RIC) for each sample

       For each sample:
         Raw Spectra and background-subtracted mass
           spectra of target  compounds identified
         Quantitation reports
         Mass Spectra of all reported TICs with three
           best library matches

    c.  Standards Data (All Instruments)
       Initial Calibration Data  (Form VI LCV-1, LCV-2, LCV-3)
       RICs  and Quantitation Reports  for all  Standards
       Continuing Calibration Data (Form VII  LCV-1,  LCV-2, LCV-3)
       RICs  and Quantitation Reports  for all  Standards

    d.  Raw QC Data
       BFB
       Blank Data
       Matrix Spike/Matrix Spike Duplicate Data
         (if  Region requests)
                                                 FORM DC-2-1                               OLC03.2

-------
             LOW CONCENTRATION WATER ORGANICS COMPLETE SDG FILE (CSF)  INVENTORY SHEET (Con't)
CASE NO.
                      SDG NO.
                                          SDG NOS. TO FOLLOW
                                          CLIENT NO. 	
                                                                                  PAGE NOs
                                                                                 FROM    TO
  CHECK
LAB    EPA
5.  Semivolatiles Data
   a. QC  Summary
       Deuterated Monitoring Compound (Form II LCSV)
       MS/MSD Summary (Form III LCSV)  (if Region requests)
       Method Blank Summary (Form IV LCSV)
       GC/MS Instrument Performance Check (Form V LCSV)
       Internal Standard Area and RT Summary (Form VIII LCSV)
   b. Sample  Data
       TCL Results - (Form I LCSV-1, LCSV-2)
       Tentatively Identified Compounds (Form I LCSV-TIC)
       Reconstructed Total Ion Chromatograms (RIC)  for each sample
       For each sample:
         Raw Spectra and background-subtracted mass
           spectra  of  target  compounds
         Quantitation reports
         Mass Spectra of TICs with three best library matches

    c.  Standards,; Data (All Instruments)
       Initial Calibration Data  (Form VI LCSV-1, LCSV-2, LCSV-3)
       RICs and Quantitation Reports for all Standards
       Continuing Calibration Data  (Form VII LCSV-1, LCSV-2, LCSV-3)
       RICs and Quantitation Reports for all Standards
    d.  Raw QC Data
       DFTPP
       Blank Data
       Matrix Spike/Matrix Spike Duplicate Data (if Region requests)
6. Pesticides Data
   a.  QC Summary
       Surrogate Percent Recovery Summary (Form II LCP)
       MS/MSD Duplicate Summary  (Form III LCP-1) (if Region requests)
      Laboratory Control Sample  Recovery  (Form III LCP-2)
       Method Blank Summary  (Form IV LCP)
    b.  Sample Data
       TCL Results - Organic Analysis Data Sheet (Form I LCP)
       Chromatograms  (Primary Column)
       Clnromatograms from second GC column confirmation
       GC Integration report or data system printout
       Manual work sheets
                                                FORM DC-2-2
                                                                                         OLC03.2

-------
             LOW CONCENTRATION WATER ORGANICS COMPLETE SDG FILE (CSF)  INVENTORY SHEET (Con't)
CASE NO.
                      SDG NO.
                                          SDG NOS. TO FOLLOW
                                          CLIENT NO. 	
                                                                                   PAGE NOS
                                                                                 FROM    TO
  CHECK
LAB    EPA
6. Pesticides Data (Con't)
   c. Standards Data
       Initial Calibration of Single Component Analytes
         (Form VI  LCP-1 and LCP-2)
       Initial Calibration of Multicomponent Analytes (Form VI LCP-3)
       Analyte Resolution Summary (Form VI LCP-4)
       Performance Evaluation Mixture  (Form VI LCP-5)
       Individual Standard Mixture A (Form VI LCP-6)
       Individual Standard Mixture B (Form VI LCP-7)
       Calibration Verification Summary  (Form VII LCP-1)
       Calibration Verification Summary  (Form VII LCP-2)
       Analytical Sequence  (Form VIII LCP)
       Florisil Cartridge Check (Form IX LCP)
       Pesticide Identification Summary for Single
         Component Analytes (Form X LCP-1)
       Pesticide Identification Summary for
         Multicomponent Analytes (Form X LCP-2)
       Chromatograms and data system printouts
         A printout of retention times and corresponding peak
         areas or peak heights
       Raw QC Data
       Blank Data
       Matrix Spike/Matrix Spike Duplicate Data  (if Region requests)
       Laboratory Control  Sample Data
    e. Raw Florisil Data
 7. Miscellaneous  Data
       Original preparation  and  analysis  forms  or  copies
         of preparation and analysis logbook pages
       Internal sample and sample  extract transfer
         chain-of-custody records
       Screening  records
       All  instrument output,  including strip charts
       from screening activities (describe or list)
                                                 FORM DC-2-3
                                                                                          OLC03.2

-------
             LOW CONCENTRATION WATER ORQANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET (Con't)
CASE NO.
                      SDG NO.
                                           SDG NOS.  TO  FOLLOW
                                          •CLIENT NO. 	
                                                                                   PAGE NOs
                                                                                 FROM    TO
                                     CHECK
                                  LAB    EPA
8.  EPA Shipping/Receiving Documents
     Airbills  (No. of shipments 	)
     Chain-of-Custody Records
     Sample Tags
     Sample Log-in Sheet  (Lab  & DC1)
     Miscellaneous Shipping/Receiving Records  (describe or list)
9. Internal Lab Sample Transfer Records and Tracking Sheets
    (describe  or list)
10 .  Other- Records
    (describe or list)
    Telephone Communication Log
11. Comments;
Completed by:
  (CLP Lab)
                               (Signature)
(Printed Name/Title)
(Date)
Verified  by:
  (CLP Lab)
                               (Signature)
(Printed Name/Title)
Audited by:
   (EPA)
                               (Signature)
(Printed Name/Title)'
(Date)
                                                FORM DC-2-4
                                                                                          OLC03.2

-------
                                        EXHIBIT C
                                TARGET COMPOUND LIST AND
                          CONTRACT REQUIRED QUANTITATION LIMITS
NOTE:  Specific quantitation limits are highly matrix-dependent.   The
       quantitation limits listed herein are provided for guidance and may not
       always be achievable.

       The CRQL values listed on the following pages are.based on the analysis
       of samples according to the specifications given in Exhibit D.
                                            C-l                                     OLC03.2

-------
                            THIS PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                    C-2

-------
       Exhibit C  - Target  Compound List  and  Contract Required Quantitation Limits

                                    Table of Contents

                                                                                     Page

1.0    VOLATILES  TARGET COMPOUND LIST AND CONTRACT REQUIRED
       QUANTITATION LIMITS   	  	 5

2.0    SEMIVOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED
       QUANTITATION LIMITS   	 .  . 7
3 .0    PESTICIDES/AROCLORS TARGET COMPOUND LIST AND CONTRACT
       REQUIRED QUANTITATION LIMITS . .  	 .... 9
                                            C-3                                    OLC03.2

-------
                            THIS PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                    C-4

-------
1.0
                                                                   Exhibit C -- Section 1
                                                                          Volatiles  (VGA)
                  VOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED
                                   QUANTITATION LIMITS
                                                                  Ouantitation Limits

1.
2.
3.
4.
5.
6.
7.
8.

9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
31.
32.
33.
34.
35.
36.
37.
38.
Volatiles
Dichlorodifluoromethane
Chi oromethane
Vinyl Chloride
Brotnomethane
Chloroethane
Trichlorofluoromethane
1 , 1 -Dichloroethene
1,1,2-Trichloro-
1,2 , 2-trif luoroethane
Acetone
Carbon Disulfide
Methyl Acetate
Methylene Chloride
trans-1, 2 -Dichloroethene
Methyl tert -Butyl Ether
1 , 1 -Dichloroethane
cis-1, 2 -Dichloroethene
2-Butanone
Bromochloromethane
Chloroform
1,1, 1-Trichloroethane
Cyclohexane
Carbon Tetrachloride
Benzene
1 , 2 -Dichloroethane
Trichloroethene
Me t hy 1 eye 1 ohexane
1 , 2 -Dichloropropane
Bromodichloromethane
cis - 1 , 3 -Dichloropropene
4 -Methyl -2 -pentanone
Toluene
trans-1, 3 -Dichloropropene
1,1,2 -Trichloroethane
Tetrachloroethene
2-Hexanone
Dibromochloromethane
1 , 2 -Dibromoethane
Chlorobenzene
CAS Number
75-71-8
74-87-3
75-01-4
74-83-9
75-00-3
75-69-4
75-35-4

76-13-1
67-64-1
75-15-0
79-20-9
75-09-2
156-60-5
1634-04-4
75-34-3
156-59-2
78-93-3
74-97-5
67-66-3
71-55-6
110-82-7
56-23-5
71-43-2
107-06-2
79-01-6
108-87-2
78-87-5
75-27-4
10061-01-5
108-10-1
108-88-3
10061-02-6
79-00-5
127-18-4
591-78-6
124-48-1
106-93-4
108-90-7
Water
ug/L
0.50 .
0.50
0.50
0.50
0.50
0.50
0.50

0.50
5.0
0.50
0.50
0.50
0.50
, 0.50
0.50
0.50
5.0
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
0.50
5.0
0.50
0.50
0.50
0.50
5.0
0.50
0.50
0.50
                                            C-5
                                                                                   OLC03.2

-------
Exhibit C -- Section  1
Volatiles  (VOA)  (Con't)
1.0
                   VOLATILE TARGET COMPOUND LIST AND CONTRACT REQUIRED
                                QUANTITATION LIMITS  (Con't)
                                                                  Ouantitation Limits

39.
40.
41.
42.
43.
44.
45.
46.
47.
48.
49.
50.
Volatiles
Ethylbenzene
Xylenes (total)
Styrene
Bromoform
I sopropylbenzene
1,1,2,2 -Tetrachloroethane
i , 3 -Dichlorobenzene
1 , 4 -Dichlorobenzene
1 , 2 -Dichlorobenzene
1 , 2 -Dibromo- 3 -chloropropane
1,2,4 -Trichlorobenzene
1,2,3 -Trichlorobenzene
CAS Number
100-41-4
1330-20-7
100-42-5
75-25-2
98-82-8
79-34-5
541-73-1
106-46-7
95-50-1
96-12-8
120-82-1
87-61-6
Water
0,
0,
0,
0,
0,
0,
0,
0,
0.
0,
0,
0,
.50
.50
.50
.50
.50
.50
.50
.50
.50
.50
.50
.50
OLC03.2
                                            C-6

-------
                                                                   Exhibit C -- Section 2
                                                                     Semivolatiles  (SVGA)
2.0
SEMIVOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED
                   QUANTITATION LIMITS
                                                               Ouantitation Limits

51.
52.
53.
54.
55.
56.
57.
58.
59.
60.
61.
62.
63.
64.
65.
66.
67.
68.
69.
70.
71.
72.
73.
74.
75.
76.
77.
78.
79.
80.
81.
82.
83.
84.
Semivolatiles
Benzaldehyde
Phenol
bis- (2-Chloroethyl) ether
2 - Chlorophenol
2 -Methylphenol
2,2' -oxybis (1-Chloropropane) l
Acetophenone
4 -Methylphenol
N-Nitroso-di-n-propylamine
Hexachl oroethane
Nitrobenzene
Isophorone
2 -Nitrophenol
2 , 4 -Dimethylphenol
bis (2-Chloroethoxy)methane
2 , 4 -Dichlorophenol
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
Caprolactam
4 -Chloro- 3 -methylphenol
2 -Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6-Trichlorophenol
2,4,5 -Tr ichlorophenol
1,1' -Biphenyl
2 - Chloronaphthalene
2-Nitroaniline
Dimethylphthalate
2, 6-Dinitrotoluene
Acenaphthylene
3-Nitroaniline
Acenaphthene
2 , 4 -Dinitrophenol
CAS
Number
100-52-7
108-95-2
111-44-4
95-57-8
95-48-7
108-60-1
98-86-2
106-44-5
621-64-7
67-72-1
98-95-3
78-59-1
88-75-5
105-67-9
111-91-1
120-83-2
91-20-3
106-47-8
87-68-3
105-60-2
59-50-7
91-57-6
77-47-4
88-06-2
95-95-4
92-52-4
91-58-7
88-74-4
131-11-3
606-20-2
208-96-8
99-09-2
83-32-9
51-28-5
Water
• g/L
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0.
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
20
5.0
5.0
20
5.0
5.0
5.0
20
5.0
20
        Previously known by the name bis(2-Chloroisopropyl)ether.

                                            C-7
                                                                                   OLC03.2

-------
Exhibit C -- Section 2
Semivolatiles  (SVOA)  (Con't)
2.0
                SEMIVOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED
                               QUANTITATION LIMITS  (Con't)
                                                               Ouantitation Limits

85.
86.
87.
88.
89.
90.
91.
92.
93.
94.
95.
96.
97.
98.
99.
100.
101.
102.
103.
104.
105.
106.
107.
108.
109.
110.
111.
112.
113.
114.
115.
Semivolatiles
4-Nitrophenol
Dibenzofuran
2 , 4 -Dinitrotoluene
Diethylphthalate
Fluorene
4 - Chlorophenyl -phenylether
4 -Nitroaniline
4, 6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine
1,2,4,5 Tetrachlorobenzene
4 -Bromophenyl -phenylether
Hexachlorobenzene
Atrazine
Pentachlorophenol
Phenanthrene
Anthracene
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Benzo (a) anthracene
Chrysene
bis(2-Ethylhexyl)phthalate ,
Di-n-octylphthalate
Benzo (b) fluoranthene
Benzo (k) fluoranthene
Benzo (a) pyrene
Indeno ( 1 , 2 , 3 - cd ) pyrene
Dibenzo (a, h) anthracene
Benzo (g, h, i) perylene
CAS
Number
100-02-7
132-64-9
121-14-2
84-66-2
86-73-7
7005-72-3
100-01-6
534-52-1
86-30-6
95-94-3
101-55-3
118-74-1
1912-24-9
87-86-5
85-01-8
120-12-7
84-74-2
206-44-0
129-00-0
85-68-7
91-94-1
56-55-3
218-01-9
117-81-7
117-84-0
205-99-2
207-08-9
50-32-8
193-39-5
53-70-3
191-24-2
Water
•g/L
20
5.0
5.0
5.0
5.0
5.0
20
20
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
5.0
OLC03.2
                                            C-8

-------
3.0
                                                                    Exhibit C  -  Section 3
                                                           Pesticides/Aroclors (PEST/ARO)
PESTICIDES/AROCLORS TARGET COMPOUND LIST AND CONTRACT
             REQUIRED QUANTITATION LIMITS
                                                               Ouantitation Limits

116.
117.
118.
119.
120.
121.
122.
123.
124.
125.
126.
127.
128.
129.
130.
131.
132.
133.
134.
135.
136.
137.
138.
139.
140.
141.
142.
143.
Pesticides/Aroclors
alpha -BHC
beta-BHC
delta-BHC
gamma-BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide2
Endosulfan I
Dieldrin
4,4' -DDE
Endrin
Endosulfan II
4,4' -ODD
Endosulfan sulfate
4,4' -DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha - Chi ordane
gamma - Chlordane
Toxaphene
Aroclor-1016
Aroclor-1221
Aroclor-1232
Aroclor-1242
Aroclor-1248
Aroclor-1254
Aroclor-1260
CAS Number
319-84-6
319-85-7
319-86-8
58-89-9
76-44-8
309-00-2
1024-57-3
959-98-8
60-57-1
72-55-9
72-20-8
33213-65-9
72-54-8
1031-07-8
50-29-3
72-43-5
53494-70-5
7421-93-4
5103-71-9
5103-74-2
8001-35-2
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
Water
• g/L
0.01
0.01
0.01
0.01
0.01
0.01
0.01
0.01
0.02
0.02
0.02
0.02
0.02
0.02
0.02
0.10
0.02
0.02
0.01
0.01
1.0
0.20
0.40
0.20
0.20
0.20
0.20
0.20
        2Only the exo-epoxy isomer (isomer  B)  of  heptachlor epoxide is reported
 on the data reporting forms  (Exhibit B).

                                            C-9
                                                                                   OLC03.2

-------
                       EXHIBIT D
METHOD FOR THE ANALYSIS OF LOW CONCENTRATION WATER FOR
        VOLATILE  (PURGEABLE) ORGANIC COMPOUNDS
                        D-l/VOA                                  OLC03.2

-------
                           THIS PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                  D-2/VOA

-------
                      Exhibit D -- Analytical Methods for Volatiles

                                    Table of Contents

                                                                                     Page


1.0    SCOPE AND APPLICATION	5

2.0    SUMMARY OF METHOD	6

3.0    DEFINITIONS  	 6

4.0    INTERFERENCES	 7

5 .0    SAFETY	8

6.0    EQUIPMENT AND SUPPLIES 	 9
       6.1    Glassware	9
       6.2    pH Paper	9
       6.3    Balances   	9
       6.4    Purge and  Trap Device	'	.  . 9
       6.5    Gas Chromatograph/Mass Spectrometer (GC/MS)  System  .  .  .  . ;	11

7.0    REAGENTS AND STANDARDS	15
       7.1    Reagents   	15
       7.2    Standards	15

8.0    SAMPLE COLLECTION, PRESERVATION, AND STORAGE  	  20
       .8.1    Sample Collection and Preservation  	  20
       8.2    Procedure  for Sample Storage   	  20
       8.3    Temperature Records for Sample Storage  	  20
       8.4    Contract Required Holding Times  	  	  21

9.0    CALIBRATION AND STANDARDIZATION	'	21
       9.1    Instrument Operating Conditions  	  21
       9.2    Instrument Performance Check -- 4-Bromofluorobenzene (BFB)   	  23
       9.3    Initial Calibration  	  24
       9.4    Continuing Calibration   	  	  28

10.0   SAMPLE ANALYSIS	31
       10.1   Summary of Sample Analysis   	31
       10.2   Procedure for Sample Analysis  	  31

11.0   DATA ANALYSIS AND CALCULATIONS	35
       11.1   Qualitative Identification of Target Compounds   	  35
       11.2   Qualitative Identification of Non-Target Compounds  	  36
       11.3   Calculations   	  38
       11.4   Technical Acceptance Criteria for Sample Analysis  	  41
       11.5   Corrective Action for Sample Analysis  	  42

12.0   QUALITY CONTROL  (QC)	44
       12.1   Blank Analyses   	44
       12.2   Matrix Spike/Matrix Spike Duplicate (MS/MSD)    	  47
       12.3   Method Detection Limit (MDL) Determination   	'  	  49
                                         D-3/VOA                                   OLC03.2

-------
                      Exhibit D -- Analytical Methods for Volatiles





                              Table of Contents  (Continued)





                                                                                     Page








13.0   METHOD PERFORMANCE	:   51





14.0   POLLUTION PREVENTION	   51





15.0   WASTE MANAGEMENT  	   51





16.0   REFERENCES	   51





17.0   TABLES/DIAGRAMS/FLOWCHARTS  ..."	   52
 OLC03.2                                  D-4/VOA

-------
1.0
                                                         exhibit D Volatiies -- Section 1
                                                                    Scope and Application
       SCOPE AND APPLICATION
1.1    The analytical method that follows is designed to analyze water samples
       containing low concentrations of the volatile compounds listed in the
       Target Compound List (TCL) in Exhibit C.  The majority of the samples
       are expected to be obtained from drinking water and well/ground water
       type sources around Superfund sites.   The method is based on EPA Method
       524.2 and the volatile method contained in the Contract Laboratory
       Program (CLP)  Statement of Work (SOW),  "Organic Analysis, Multi-Media,
       Multi-Concentration".   The sample preparation and analysis procedures
       included in this method are based on purge and trap Gas
       Chromatograph/Mass Spectrometer (GC/MS) techniques.1

1.2    Dichlorodifluoromethane,  Trichlorofluoromethane, l,l,2-Trichloro-l,2,2T
       trifluoroethane, Methyl Acetate, 1,2,3-Trichlorobenzene,  Methyl tert-
       butyl ether, Cyclohexane,  Methylcyclohexane, and Isopropylbenzene have
       been added to the TCL.

1.3    Problems that have been associated with the following compounds analyzed
       by this method include:

       •    Chloromethane,  vinyl  chloride, bromotnethane,  and chloroethane may
            display peak broadening if  the compounds  are  not delivered to the  GC
            column  in a tight  band.

       •    Acetone,  hexanone, 2-butanone, and 4-methyl-2-pentanone  have poor
            purge efficiencies.

       •    1,1,1-Trichloroethane and all of the dichloroethanes may
            dehydrohalogenate  during storage or analysis.

       •    Tetrachloroethane  and 1,1-dichloroethane  may  be  degraded by
            contaminated transfer lines in purge and  trap systems and/or active
            sites in  trapping  materials.

       •    Chloromethane may be  lost if  the purge flow is too fast.

       •    Bromoform is one  of the compounds  most likely to be  adversely
            affected  by cold  spots and/or active sites in the transfer lines.
            Response  of its quantitation  ion (m/z 173)  is directly affected  by
            the tuning of 4-Bromofluorobenzene (BFB)  at ions m/z 174/176.
            Increasing the m/z 174/176  ratio within the specified Quality
            Control (QC)  limits may improve  bromoform response.

       •    Due to  the lower  quantitation limits required by this method, extra
            caution must be exercised when identifying compounds.
       '•This analytical method includes the use of Deuterated Monitoring
Compounds  (DMC) for precision and accuracy assessment.

                                         D-5/VOA
OLC03.2

-------
Exhibit D Volatiles — Sections 2 & 3
Summary of Method

2.0    SUMMARY OF METHOD

2.1    An inert gas is bubbled through a 25 milliliter (tnL) sample contained in
       a specially designed purging chamber at ambient temperature causing the
       purgeables to be transferred from the water/aqueous phase to the vapor
       phase.  The vapor is swept through a sorbent column where the purgeables
       are trapped.  After purging is completed, the sorbent column is heated
       and backflushed with the inert gas to desorb the purgeables onto a Gas
       Chromatograph  (GC) wide-bore capillary column.  The GC is temperature
       programmed to separate the purgeables, which are then detected with a
       Mass Spectrometer (MS).

2.2    Deuterated Monitoring Compounds  (DMCs) and internal standards are added
       to all samples and blanks.  The target compounds and DMCs are identified
       in the samples and blanks by analyzing standards that contain all target
       compounds, DMCs, and internal standards under the same conditions and
       comparing resultant mass spectra and GC retention times.  A Relative
       Response Factor  (RRF) is established for each target compound and DMC
       during the initial and continuing calibrations.  The mass spectra
       response from the Extracted Ion Current Profile (EICP) for the primary
       quantitation ion produced by that compound is compared to the mass
       spectra response for the primary quantitation ion produced by the
       associated internal standard compound.  Each identified target compound
       and DMC is quantitated by comparing the instrument  response for the
       compound in the sample or blank with the instrument response of the
       associated internal standard, while taking into account the RRF from the
       most  recent mid-point calibration, the sample volume, and any sample
       dilutions.

2.3    Non-target compounds are identified by comparing the resultant mass
       spectra from the non-target compounds to mass spectra contained in the
       NIST/EPA/NIH  (May 1992 release or later) and/or Wiley  (1991 release or
       later) or equivalent mass spectral library.  Non-target compounds are
       quantitated by comparing the mass spectra response  from the total ion
       chromatograms  to  the mass spectra response of the nearest internal
       standard compound.  A RRF of 1 is assumed.

3.0    DEFINITIONS

       See Exhibit G  for a complete list of  definitions.
 OLC03.2                                  D-6/VOA

-------
                                                         Exhibit D Volatiles -- Section 4
                                                                            Interferences
4 .0    INTERFERENCES
4.1    Method interference may be caused by impurities in the purge gas,
       organic compounds out-'gassing from the plumbing ahead of the trap,  and "
       solvent vapors in the laboratory.  The analytical system must be
       demonstrated to be free from contamination under the conditions of  the
       analysis by running laboratory method and instrument blanks as described
       in Section 12.   The use of non-polytetrafluorbethylene (PTFE) tubing,
       non-PTFE thread sealants, or flow controllers with rubber components in
       the purging device should be avoided.

4.2    Samples can be contaminated by diffusion of purgeable organics
       (particularly methylene chloride, fluorocarbons, and other common
       laboratory solvents)  through the septum seal into the sample during
       storage and handling.  Therefore, these samples must be stored
       separately from other laboratory samples and standards and must be
       analyzed in a room whose atmosphere is demonstrated to be free of all
       potential contaminants which will interfere with the analysis.

4.3    Contamination by carryover can occur whenever high level and low level
       samples are sequentially analyzed.  To reduce carryover,  the purging
       device and sampling syringe must be rinsed with reagent water between
       sample analyses.  Whenever an unusually concentrated sample is
       encountered, it must either be followed by analysis of an instrument
       blank, or the next sample must be closely monitored to check for cross
       contamination.   For samples containing large amounts of water-soluble
       materials, suspended solids, high boiling compounds, or high purgeable
       levels, it may be necessary to wash out the purging device with a
       detergent solution between analyses, rinse it with distilled water, and
       then dry it in an oven-at 105°C.  The trap and other parts of the system
       are also subject to contamination; therefore, frequent bakeout and
       purging of the entire system may be required.

4.4    The laboratory where volatile analysis is performed should be completely
       free of solvents.
                                         D-7/VOA                                  OLC03.2

-------
Exhibit D Volatiles -- Section 5
Safety

5.0    SAFETY

5.1    The toxicity or carcinogenicity of each reagent used in this method has
       not been precisely defined; however, each chemical compound should be
       treated as a potential health hazard.  From this viewpoint, exposure to
       these chemicals must be reduced to the lowest possible level by whatever
       means available.   The laboratory is responsible for maintaining a
       current awareness file of OSHA regulations regarding the safe handling
       of the chemicals specified in this method.  A reference file of material
       data handling sheets should also be made available to all personnel
       involved in the chemical analysis.

5.2    The following analytes covered by this method have been tentatively
       classified as known or suspected, human or mammalian carcinogens:
       benzene; carbon tetrachloride; chloroform; and vinyl chloride.  Primary
       standards of these toxic compounds should be prepared in a hood.  A
       NIOSH/MESA-approved toxic gas respirator should be worn when the analyst
       handles high concentrations of these toxic compounds.
 OLC03.2                                  D-8/VOA

-------
                                                         Exhibit D Volatiles -- Section 6
                                                                   Equipment and Supplies
6.0
6.1

6.1.1



6.1.2


6.1.3

6.1.4

6.1.5

6.1.6
       EQUIPMENT AND SUPPLIES
Brand names, suppliers, and part numbers are for illustrative purposes
only.  No endorsement is implied:- Equivalent-performance may be     -
achieved using equipment and supplies other than those specified here,
but demonstration of equivalent performance meeting the requirements of
the Statement of Work  (SOW) is the responsibility of the Contractor. The
Contractor shall document any use of alternate equipment or supplies in
the Sample Delivery Group  (SDG) Narrative.

Glassware

    Syringes  -  25  milliliters  (mL),  gas-tight with  shut-off  valve.   Micro
    syringes  -  10  microliters  (• L) and larger,  0.006 inch (0.15 mm) ID
    needle.

    Syringe  Valve  -  two-way, with Luer ends (three  each),  if applicable
    to the purging device.

    Pasteur  Pipets - disposable.

    Vials and Caps - assorted  sizes.

    Volumetric Flasks,  class A with ground-glass  stoppers.

    Bottles  - 15 mL, screw-cap,  with PTFE cap liner.
6.2    pH Paper - wide range

6.3    Balances

       Balances must be analytical and capable of accurately weighing ±0.0001
       g.  The balance must be calibrated witH class S weights or known
       reference weights once per'each 12-hour work shift.  The balance must be
       calibrated with class S weights at a minimum of once per month.  The
       balance must also be annually checked by a certified technician.

6 .4    Purge and Trap Device

       The purge and trap device  consists of three separate pieces of
       equipment: the sample purge chamber, trap, and the desorber.  Several
       complete devices are now commercially available.

6.4.1      The  sample purge  chamber must  be  designed  to accept  25 mL  samples
           with a water  column at  least 10 centimeters  (cm)  deep.   The  gaseous
           head space between the  water column  and the  trap  must have a total
           volume of less  than 15  mL.  The purge  gas  must  pass  through  the water
           column as finely  divided bubbles,  each with  a diameter of  less than  3
           millimeters  (mm)  at the origin.   The purge gas  must  be introduced no
           more than 5 mm  from the base of the  water  column.

6.4.2      The  trap must be  at least  25 cm long and have an  inside  diameter of
           at least 0.105  inch (2.667 mm).   The trap  must  be packed to  contain
           the  following minimum lengths  of  absorbents:  (starting from  inlet)
           0.5  cm silanized  glass  wool, 1 cm methyl silicone, 8 cm  of 2,6-

                                         D-9/VOA
                                                                            OLC03.2

-------
Exhibit D Volatiles -- Section 6
Equipment and Supplies  (Con't)

          diphenylene  oxide  polymer (Tenax-GC,  60/80  mesh),8  cm of  silica gel
           (Davison Chemical,  35/60 mesh,  grade  15  or  equivalent),  7  cm of
          coconut  charcoal  (prepare from Barnebey  Cheney,  CA-580-26,  or
          equivalent,  by crushing through 26  mesh  screen),  and 0.5 cm silanized
          glass  wool.   A description of  the trap used for  analysis shall be
          provided in  the SDG Narrative.

6.4.3     The  desorber must  be capable of rapidly  heating  the  trap to 180°C.
          The  polymer  section of the trap should not  be heated higher than
          180°C  and the remaining sections should  not exceed 220°C during
          bakeout  mode.

6.4.4     Trap Packing

6.4.4.1       2,6-Diphenylene oxide polymer,  60/80 mesh chromatographic  grade
               (Tenax GC or equivalent).

6.4.4.2       Methyl silicone packing, 3.0 percent OV-l on  Chromosorb W, 60/80
              mesh  (or equivalent).

6.4.4.3       Silica gel, 35/60 mesh, Davison, grade  15  (or equivalent).

6.4.4.4       Coconut  charcoal  (prepare  from  Barnebey Cheney,  CA-580-26, or
              equivalent, by  crushing through 26 mesh screen).

6.4.4.5       Alternate  sorbent  traps may be  used  if:

               •    The trap packing materials do not introduce contaminants which
                   interfere with identification and quantitation of  the
                   compounds listed in Exhibit C (Volatiles);

               •    The analytical results generated using the trap meet the
                   initial and continuing calibration technical acceptance
                   criteria listed in the SOW and the Contract Required
                   Quantitation Limits (CRQLs) listed in Exhibit C (Volatiles);
                   or

               •    The trap can accept up to  1000 nanograms  (ng) of each compound
                   listed in Exhibit C (Volatiles)  without becoming overloaded.

6.4.4.5.1        The alternate trap must be designed to optimize performance.
                  Follow the manufacturer's instructions for the use  of its
                  product.  Before use of any trap other than the one specified
                  in Section 6.4.2, the Contractor must first meet the criteria
                  listed in Section 6.4.4.5.  Once this has been demonstrated,
                  the Contractor must document its use in each SDG Narrative by
                  specifying the trap composition  (packing material/brand name,
                  amount of packing material).  Other sorbent traps include, but
                  are not limited to, Tenax/Silica Gel/Carbon Trap from EPA
                  Method 524.2, Tenax - GC/Graphpac-D Trap  (Alltech)  or  .
                  equivalent, and Vocarb 4000 Trap (Supelco) or equivalent.

6.4.4.5.2        The Contractor must maintain documentation that the alternate
                  trap meets the criteria listed in Section 6.4.4.5.   The minimum
                  documentation requirements are as follows:

OLC03.2                                  D-10/VOA

-------
                                                         Exhibit D Volatiles -- Section 6
                                                           Equipment and Supplies  (Con't)

6.4.4.5.2.1          Manufacturer-provided information concerning the performance
                     characteristics  of the trap.

6.4.4.5.2.2          Reconstructed ion chromatograms and data system reports
                     generated on the Contractor's GC/MS used for CLP analyses:

                     •   From instrument blank analyses which demonstrate that
                        there are no contaminants which interfere with the
                        volatile analysis when using the alternate trap; and

                     •   From initial and continuing calibration standards
                        analyzed using the trap specified in Section 6.4.4.

6.4.4.5.2.3          Based on Contractor-generated data described above,  the
                     Contractor must  complete a written comparison/review,  which
                     has been signed  by the Laboratory Manager,  certifying that:

                     •   The alternate trap performance meets the technical
                        acceptance criteria listed in Sections 9.3.5 and 9.4.5;

                     •   The low point initial calibration standard analysis has
                        adequate sensitivity to meet the volatile CRQLs,-

                     •   The high point initial calibration standard analysis was
                        not overloaded; and

                     •   The alternate trap materials do not introduce
                        contaminants which interfere with the identification
                        and/or quantitation of the compounds listed in Exhibit  C
                         (Volatiles).

6.4.4.5.2.4          The documentation must be made available to USEPA  during on-
                     site laboratory evaluations or sent to USEPA upon  request of
                     the USEPA CLP Project Officer (CLP PO)  or the Organic
                     Program Manager  at the Analytical Operations/Data  Quality
                     Center (AOC).

6.4.5      The purge  and trap  apparatus may be  assembled as a  separate unit  or
           be an  integral  unit  coupled with a Gas  Chromatograph  (GC).

6.5    Gas Chromatograph/Mass Spectrometer (GC/MS) System

6.5.1      Gas Chromatograph -  The  GC system must  be capable  of  temperature
           programming  and have a  flow controller  that maintains  a constant
           column flow  rate throughout desorption  and temperature program
           operations.   The system  must  include or be interfaced  to  a purge  and
           trap system  as  specified in Section  6.4 and have all required
           accessories  including syringes,  analytical columns, and gases.  All
           GC carrier gas  lines must  be constructed from stainless steel  or
           copper tubing.   Non-polytetrafluoroethylene  (PTFE)  thread sealants,
           or flow controllers  with rubber components, are not to be used.   The
           column oven  must be  cooled to  10°C if adequate separation of gaseous
           compounds  is not achieved  (Section 9.1.2.3);  therefore, a subambient
           oven controller is  required.
                                         D-ll/VOA                                 OLC03.2

-------
Exhibit D Volatiles -- Section 6
Equipment and Supplies (Con't)

6.5.2     Gas Chromatography Columns

          A  description  of  the column used  for  analysis  shall  be provided  in
          the SDG Narrative.

6.5.2.1       Minimum length 30 m x 0.53 mm ID VOCOL (Supelco)  or equivalent
              fused silica widebore capillary column with 3 micrometers (vim)
              film thickness.

6.5.2.2       Minimum length 30 m x 0.53 mm ID DB-624  (J & W Scientific) or
              equivalent fused silica widebore capillary column with 3 urn film
              thickness.

6.5.2.3       Minimum length 30 m x 0.53 mm ID AT-624  (Alltech) or equivalent
              fused silica widebore capillary column with 3 urn film thickness.

6.5.2.4       Minimum length 30 m x 0.53 mm ID HP-624  (Hewlett-Packard) or
              equivalent fused silica widebore capillary column with 3 um film
              thickness.

6.5.2.5       Minimum length 30 m x 0.53 mm ID RTx-624 (Restek) or equivalent
              fused silica widebore capillary column with 3 vim film thickness.

6.5.2.6       Minimum length 30 m x 0.53 mm ID BPX-624 (SGE) or equivalent  fused
              silica widebore capillary column with 3 um film thickness.

6.5.2.7       Minimum length 30 m x 0.53 mm ID CP-Sil  13CB  (Chrompack) or
              equivalent fused silica widebore capillary column with 3 vim  film
              thickness.

6.5.3      A capillary column is considered equivalent if:

           •    The column does not  introduce contaminants which interfere  with
                the identification and quantitation of  the compounds listed in
                Exhibit C (Volatiles);

           •    The analytical  results generated using  the column meet  the
                initial and  continuing calibration technical acceptance criteria
                listed in the SOW, and the CRQLs  listed in Exhibit  C  (Volatiles);

           •    The column can accept up to  1000  ng of  each  compound listed in
                Exhibit C (Volatiles) without becoming  overloaded;  and

           •    The column provides  equal or better resolution  of the compounds
                listed in Exhibit  C  (Volatiles)  than  the  columns listed in
                Section 6.5.2.

 6.5.3.1      As applicable,  follow the manufacturer's instructions for use of
              its product.

 6.5.3.2      The Contractor must maintain  documentation that  the  column met the
              criteria in Section 6.5.3.  The minimum documentation is as
              follows:
 OLC03.2                                  D-12/VOA

-------
                                                         Exhibit D Volatiles -- Section 6
                                                           Equipment and Supplies (Con't)

6.5.3.2.1        Manufacturer provided information concerning the performance
                 characteristics of the column.

6.5.3.2.2        Reconstructed ion chromatograms and data system reports
                 generated on the GC/MS used for the CLP analyses:

                 •    From instrument  blanks which demonstrate that  there are no
                      contaminants which interfere with the  volatile analysis
                      when using the column; and

                 •    From initial and continuing calibration standards analyzed
                      using the alternate column.

6.5.3.5       Based on the Contractor-generated data described above,  the
              Contractor shall complete a written comparison/review,  signed by
              the Laboratory Manager,  certifying that:

              •    The  alternate column performance meets the  technical
                  acceptance  criteria in  Sections  9.3.5 and 9.4.5;

              •    The  low point initial calibration  standard  analysis has
                  adequate  sensitivity to meet  the volatile CRQLs;

              •    The  high  point  initial  calibration standard analysis was not
                  overloaded;  and

              •    The  column  does not introduce contaminants  which  interfere
                  with the  identification and/or quantitation of compounds
                  listed  in Exhibit C  (Volatiles).

6.5.3.6       The documentation must be made available  to USEPA during on-site
              laboratory evaluations or sent to USEPA upon request of the CLP PO
              or the Organic Program Manager at the AOC.

6.5.4     PACKED COLUMNS  CANNOT BE USED.

6.5.5     Mass  Spectrometer (MS)  - The MS must  be  capable of  scanning from 35
          to  300 atomic mass  unit  (amu) every 2 seconds or less, utilizing 70
          volts (nominal) electron energy in the electron impact ionization
          mode  and  producing  a mass spectrum which meets all  the 4-Bromo-
          fluorobenzene (BFB)  GC/MS performance check  technical acceptance
          criteria  in  Table D-l when 50 ng  of BFB  are  injected through the GC
          inlet.  .

          NOTE:  To  ensure sufficient precision  of  mass spectral data, the MS
          scan  rate must  allow acquisition  of at least five spectra while a
          sample compound elutes  from the GC.   The purge and  trap GC/MS  system
          must  be in a room whose atmosphere is demonstrated  to be  free  of all
          potential contaminants  which will interfere  with the analysis.  The
          instrument must be  vented to outside  the facility or to a trapping
          system which prevents the release of  contaminants into the instrument
          room.

6.5.6     GC/MS Interface - Any GC/MS interface may  be used that gives
          acceptable calibration  points at  25 ng or  less per  injection for each

                                         D-13/VGA                                 OLC03.2

-------
Exhibit D Volatiles -- Section 6
Equipment and Supplies (Con't)

          of the purgeable target compounds and deuterated monitoring compounds
          and achieves all acceptable performance criteria.  GC/MS interfaces
          constructed of all-glass or glass-lined materials are recommended.
          Glass can be deactivated by silanizing with dichlorodimethylsilane.

6.5.7     Data System - A computer system must be interfaced to the MS  that
          allows the continuous  acquisition and storage on machine-readable
          media of all mass  spectra obtained throughout the duration of the
          chromatographic program.  The computer must have software that allows
          searching of any GC/MS data file for ions of a  specified mass and
          plotting such ion  abundances versus time or scan number.  This type
          of plot is defined as  an Extracted Ion Current  Profile  (EICP).
          Software must also be  available that allows integrating the abundance
          in any EICP between specified time or scan number limits.  Also, for
          the non-target compounds, software must be available that allows
          comparing sample spectra against reference library spectra.   The
          NIST/EPA/NIH  (May  1992 release or later) and/or Wiley  (1991 release
          or later) or equivalent mass spectral library shall be used as the
          reference library.   The data system must be capable of flagging all
          data files  that have been edited manually by laboratory personnel.

6.5.8     Magnetic-Tape Storage  Device - Capable of recording data and  must be.
          suitable for long-term, off-line storage.
 OLC03.2                                 D-14/VOA

-------
                                                         Exhibit D Volatiles -- Section 7
                                                                   Reagents and Standards
7.0
       REAGENTS AND STANDARDS
7.1    Reagents

       Reagents shall be dated with the receipt date and used on a first-in,
       first-out basis.  The purity of the reagents shall be verified before
       use.

7.1.1     Reagent water  - Reagent water  is defined as water in which no
          purgeable  target compound is observed at or above the Contract
          Required Quantitation Limit  (CRQL) listed in Exhibit C  for that
          compound and in which no non-target  compound is observed at or above
          2.0 micrograms per  liter  (ug/L).

7.1.1.1       Reagent water may be generated by passing tap water through a
              carbon filter bed containing about 453 g (1 Ib.)  of activated
              carbon (Calgon Corp.,  Filtrasorb-300,  or equivalent).

7.1.1.2       Reagent water may be generated using a water purification system
              (Millipore Super-Q or equivalent).

7.1.1.3       Reagent water may be prepared by boiling water for 15  minutes.
              Subsequently,  while maintaining the temperature at 90°C,  bubble a
              contaminant-free inert gas through the water for one hour.   While
              still hot,  transfer the water to a narrow-mouth screw-cap bottle,
              seal with a PTFE-lined septum,  and cap.

7.1.2     Methanol  - HPLC quality or  equivalent  --  Each  lot of methanol used
          for  analysis under  this contract must be  purged with nitrogen and
          must  be demonstrated to be  free of contaminants that interfere with
          the  measurement of  purgeable  compounds  listed  in Exhibit C.

7.2    Standards
7.2.1
       The Contractor must provide all standards to be used with this contract.
       These standards may be used only after they have been certified
       according to the procedure in Exhibit E.  The Contractor must be able to
       verify that the standards are certified.  Manufacturer's certificates of
       analysis must be retained by the Contractor and presented upon request.

           Stock Standard  Solutions
           Stock standard solutions may be  purchased or may be prepared in
           methanol  from pure  standard materials.

7.2.1.1       Prepare stock standard solutions by placing about  9.8 milliliters
              (mL) of methanol into a 10.0 tnL ground-glass stoppered volumetric
              flask.  Allow the flask to stand, unstoppered,  for about 10
              minutes or until all alcohol wetted surfaces have  dried. Weigh
              the flask to the nearest 0.1 mg.

7.2.1.2       Add the assayed reference material as described below.

7.2.1.2.1         If  the compound  is a liquid,  use a 100 microliters  (uL)  syringe
                  to  immediately add two or more drops of assayed reference

                                         D-15/VOA
OLC03.2

-------
Exhibit D Volatiles -- Section 7
Reagents and Standard (Con't)

                 .material to the flask, then reweigh.  The liquid must fall
                 directly into the alcohol without contacting the neck of the
                 flask.

7.2.1.2.2        If the compound is a gas at room temperature, fill a 5 mL
                 valved gas-tight syringe with the reference standard to the 5.0
                 mL mark.  Lower the needle to S millimeters  (mm) above the
                 methanol meniscus.  Slowly introduce the reference standard
                 above the surface of the liquid.  The gas will rapidly dissolve
                 in the methanol.  This may also be accomplished by using a
                 lecture bottle equipped with a Hamilton Lecture Bottle Septum
                  (#86600) .   Attach PTFE tubing to the side-arm relief valve and
                 direct a gentle stream of the reference standard into the
                 methanol meniscus.

7.2.1.3       Reweigh, dilute to volume,  stopper,  then mix by inverting the
              flask several times.   For non-gaseous compounds, calculate the
              concentration in ug/uL from the net gain in weight.  When compound
              purity is assayed to be 97 percent or greater,  the weight may be
              used without correction to calculate the concentration of the
              stock standard.  If the compound purity is assayed to be less than
              97 percent, the weight must be corrected when calculating the
              concentration of the stock solution.  For gaseous compounds,
              calculate the concentration in ug/uL using the Ideal.Gas Law,
              taking into account the temperature and pressure conditions within
              the laboratory.

7.2.1.4       Prepare fresh  stock standards every two months for gases or for
              reactive compounds such as styrene.  All other stock standards for
              non-gases/non-reactive purgeable compounds must be replaced after
              six months, or sooner if standard has degraded or evaporated.

7.2.2      Secondary Dilution Standards

7.2.2.1       Using stock standard solutions, prepare secondary dilution
              standards in methanol that contain the compounds of interest,
              either singly  or mixed together.  Secondary dilution standard
              solutions should be prepared at concentrations that can be easily
              diluted to prepare working standard solutions.

7.2.2.2       Prepare fresh  secondary dilution standards for gases and for
              reactive compounds, such as styrene, every month or sooner if-
              standard has degraded or evaporated.  Secondary dilution standards
              for the other  purgeable compounds must be replaced after six
              months, or sooner if standard has degraded or evaporated.

7.2.3      Working Standards

7.2.3.1       Instrument Performance Check Solution  (4-Bromofluorobenzene)

              Prepare a  25 nanograms per microliter  (ng/uL) solution of 4-Bromo-
              fluorobenzene  (BFB) in methanol.  Prepare fresh BFB solution every
              six months, or sooner if the solution has degraded or evaporated.
 OLC03.2                                 D-16/VOA

-------
                                                         Exhibit D Volatiles -- Section 7
                                                            Reagents and Standard  (Con't)

              NOTE:  The 25 ng/uL concentration is used with a 2 uL injection
              volume.   The laboratory may prepare a 50 ng/uL solution of BFB if
              a 1 uL injection volume is used.

7.2.3.2       Calibration Standard Solution

              Prepare the working calibration standard solution containing all
              of the purgeable target compounds in methanol (Exhibit C).  The
              concentration of the non-ketone target compounds and the
              associated Deuterated Monitoring Compounds (DMCs) must be 2.5
              ug/mL in the standard (i.e., final concentration).  The
              concentration of the ketones (acetone, butanone, 2-hexanone,
              4-methyl-2-pentanone)  and their associated DMCs must be 12.5 ug/mL
              in the standard (i.e.,  final concentration).   Prepare fresh
              working calibration standard solutions weekly, or sooner if
              solutions have degraded or evaporated.

7.2.3.3       Internal Standard Spiking Solution

              Prepare an internal standard spiking solution containing
              l,4-dichlorobenzene-d4,  chlorobenzene-d5, and 1,4-difluorobenzene
              in methanol at the concentration of 12.5 ug/mL for each internal
              standard.  Add 10 uL of this spiking solution into 25.0 mL of
              samples, blanks, requested Matrix Spike/Matrix Spike. Duplicate(s)
              (MS/MSD), and calibration standards for a concentration of 5.0
              ug/L.   Prepare fresh spiking solution weekly, or sooner if the
              solution has degraded or evaporated.

7.2.3.4       Deuterated Monitoring Compound (DMC) Spiking Solution

              Prepare a DMC spiking solution in methanol containing the
              compounds listed below at the following concentrations:

                                                            Concentration
               Compound                                           ug/mL
               Vinyl Chloride-d3                                  12.5
               Chloroethane-d5                                    12.5
               l,l-Dichloroethene-d2                              12.5
               2-Butanone-d5                                      12.5
               Chlorof orm-d                                       12 .5
               l,2-Dichloroethane-d4                              12.5
               Benzene-d6                                         12.5
               1,2-Dichloropropane-d6                             12.5
               Toluene-d8                                         12.5
               trans-l,3-Dichloropropene-d4                       12.5
               2-Hexanone-d5                                    •  12.5
               Bromoform-d                                        12.5
               l,l,2,2-Tetrachloroethane-d2                       12.5
               l,2-Dichlorobenzene-d4                             12.5

              Add 10 uL of this spiking solution into 25 mL of sample,  and blank
              for a concentration of 5.0 ug/L.   The DMC spiking solution is
              added to the working calibration standards so it is not to be
              added again when aqueous calibration standards are prepared.
                                         D-17/VOA                                 OLC03.2

-------
Exhibit D Volatiles -- Section 7
Reagents and Standard (Con't)
              Prepare fresh spiking solution weekly,  or sooner if the solution
              has degraded or evaporated.

7.2.3.5       Matrix Spiking Solution

              Prepare a spiking solution in methanol  that contains the following
              compounds at a-concentration of 12.5 ug/mL; 1,1-Dichloroethene,
              Trichloroethene, Chlorobenzene, Toluene,  and Benzene.   Prepare
              fresh spiking solution weekly, or sooner, if the solution has
              degraded or evaporated.

7.2.3.6       Aqueous Calibration Standard Solutions  -- Initial and Continuing

7.2.3.6.1         Prepare  five aqueous  initial  calibration standard  solutions
                  containing all  of the purgeable  target  compounds and the  DMCs
                  at  the following levels:  all  non-ketone target  compounds  and
                  non-ketone DMCs at 0.50,  1.0,  5.0,  10,  and 25 ug/L;  all ketones
                  and their associated  DMCs at  5.0, 10, 25, 50 and 125 ug/L.   It
                  is  required  that all  three xylene isomers (o-,  p-,  and m-
                  xylene)  be present in the calibration standards at
                  concentrations  of each isomer equal to  that  of  the other  target
                  compounds (i.e.,  0.50, 1.0, 5.0,  10,  and 25  ug/L).   The
                  internal standards are added  to  each  calibration standard
                  according to the procedure in Section 9.3.3.4.

7.2.3.6.2         Aqueous  calibration standards may be  prepared in a volumetric
                  flask or in  the syringe  used  to  inject  the standard into  the
                  purging  device.

7.2.3.6.2.1          Volumetric .flask  - Add an appropriate volume of working
                     calibration standard  solution to an aliquot of reagent water
                     in a volumetric flask.  Use a microsyringe and rapidly
                     inject the alcohol standard into the expanded area of the
                     filled volumetric flask.  Remove the needle as quickly as
                     possible  after  injection.  Bring to volume.   Mix by
                     inverting the  flask three times only.  Discard the contents
                     contained in the head of the flask.

7.2.3.6.2.2          Syringe  - Remove  the  plunger from a 25 mL syringe and close
                     the  syringe valve.  Pour reagent water into the  syringe
                     barrel to just  short  of overflowing.  Replace the syringe
                     plunger  and compress  the water.   Invert the syringe, open
                     the  syringe valve, and vent any residual air.  Adjust the
                     water volume  to 25.0  mL minus the amount of calibration
                     standard to be added.  Withdraw the plunger slightly  and add
                     an appropriate volume of working calibration standard
                     through  the valve bore  of the syringe.  Close  the valve  and
                     invert three  times.

 7.2.3.6.3         The 5.0 ug/L (25 ug/L for ketones)  aqueous calibration standard
                  solution is the continuing calibration standard.

 7.2.3.6.4         The methanol contained in each of the aqueous calibration
                  standards must not exceed 1 percent by volume.  •
 OLC03.2                                 D-18/VOA

-------
                                                         Exhibit D Volatiles -- Section 7
                                                            Reagents and Standard (Con't)
7.2.4     Ampulated Standard Extracts
          Standard solutions purchased from a chemical supply house as
          ampulated extracts in glass vials may be retained and used until the
          expiration date provided by the manufacturer.  If no manufacturer's
          expiration date is provided, the standard solutions as ampulated
          extracts may be retained and used for 2 years from the preparation
          date.  Standard solutions prepared by the Contractor which are
          immediately ampulated in glass vials may be retained for 2 years from
          preparation date.  Upon breaking the glass seal, the expiration times
          listed in Sections 7.2.1 to 7.2.3 will apply.  The Contractor is
          responsible for assuring that the integrity of the standards has not
          degraded  (Section 7.2.5.5).

7.2.5     Storage of Standards

7.2.5.1       Store  the  stock standards  in PTFE-sealed screw-cap  bottles with
              zero headspace at  -10°C to -20°C,  and protect the standards  from
              light.   Once one of  the bottles  containing  the stock  standard
              solution has been  opened,  it may be  used for no longer  than  one
              week.

7.2.5.2       Store  secondary dilution standards in PTFE-sealed screw-cap
              bottles  with minimal headspace at -10°C to  -20°C.   Protect the
              standards  from light.   The secondary dilution standards must be
              checked  frequently for signs of  degradation or evaporation,
              especially just prior to preparing the working calibration
              standards  from them.

7.2.5.3       Aqueous  standards  may be stored  up to 24 hours if held.in PTFE-
              sealed screw-cap vials with zero headspace  at 4°C  (±2°C), and
              protect, the standards from light.  If not so stored, they  must be
              discarded  after one  hour unless  they are set up to  be purged by  an
              autosampler.  When using an autosampler,  the standards  may be kept
              up to  12 hours in  purge tubes connected via the autosampler  to the
              purge  and  trap device.

7.2.5.4       Purgeable  standards  must be stored separately from  other
              standards.

7.2.5.5       The Contractor is  responsible for maintaining the integrity  of
              standard solutions and verifying prior to use.   This  means that
              standards  must be  brought  to room temperature prior to  use,
              checked for losses,  and checked  that all components have  remained
              in the solution.

7.2.6     Temperature Records for Storage of Standards

7.2.6.1       The temperature of all standard  storage refrigerators/freezers
              shall  be recorded  daily.

7.2.6.2       Temperature excursions shall be  noted and appropriate corrective
              actions  shall be taken to  correct problems,  when required.
                                         D-19/VOA                                 OLC03.2

-------
Exhibit D Volatiles -- Sections 7 & 8
Sample Collection, Preservation, and Storage

7.2.6.3       Corrective action Standard Operating Rrocedures (SOPs)  shall be
              posted on the refrigerators.

8.0    SAMPLE COLLECTION, PRESERVATION, AND STORAGE

8.1    Sample Collection and Preservation

8.1.1     Water samples  may be  collected in glass  containers having a total
          volume  of  at  least  40 milliliters  (mL) with  a  PTFE-lined septum and
          an open top screw-cap.  Headspace should be  avoided.  The specific
          requirements  for site sample collection  are  outlined by the Region.

8.1.2     The containers must be  filled in such  a  manner that no air bubbles
          pass through the sample as  the container is  being  filled.  Seal the
          vial so that  no  air bubbles are  entrapped  in it.

8.1.3     Water samples are preserved to a pH  of 2 at  the time of collection.

8.1.4     All samples must be iced  or refrigerated at  4°C (±2°C) from the time
          of collection until analysis.

8.2    Procedure  for  Sample Storage

8.2.1     The samples must be protected from  light and refrigerated at  4°C
           (±2°C)  from the  time  of receipt  until  60 days  after delivery  of a
          reconciled, complete  sample data package to  USEPA.  After 60  days,
          the samples may  be  disposed of in a  manner that complies with all
          applicable regulations.

8.2.2     The samples must be stored in an atmosphere  demonstrated to be free
          of all  potential contaminants and  in a refrigerator used only for
          storage of volatile samples received under this contract.

8.2.3     All volatile samples in an SDG must  be stored  together  in the same
          refrigerator.

8.2.4     Storage blanks shall be stored with samples  until  all  samples within
          an SDG  are analyzed.

8.2.5     Samples, sample  extracts, and standards must be stored  separately.

8.2.6     Volatile standards must be stored separately from semivolatile and
          pesticide/Aroclor standards.

8 .3    Temperature Records for  Sample Storage

8.3.1      The temperature  of all sample storage refrigerators  shall be  recorded
           daily.

8.3.2      Temperature excursions shall be noted and appropriate  corrective
           actions shall be taken to correct  problems,  when required.

8.3.3      Corrective action Standard Operating Procedures (SOPs)  shall  be
           posted on the refrigerators.
 OLC03.2                                 D-20/VOA

-------
                                                    exhibit D Voiatiles -- Sections 8 & 9
                                                          Calibration and Standardization
8.4    Contract Required Holding Times
       Analysis of water samples must be completed within 10 days of Validated
       Time of Sample Receipt (VTSR).   As part of USEPA's Quality Assurance
       (QA) program, USEPA may provide Performance Evaluation (PE) Samples as
       standard extracts which the Contractor is required to prepare per the
       instructions provided by USEPA.  PE samples must be prepared and
       analyzed concurrently with the samples in the SDG.  The contract-
       required 10-day holding time does not apply to PE Samples received as
       standard extracts.

9.0    CALIBRATION AND STANDARDIZATION

9.1    Instrument Operating Conditions

9.1.1      Purge  and Trap

9.1.1.1       The following are the recommended purge and trap analytical
              conditions.   The conditions are recommended unless otherwise
              noted:

               Purge Conditions
               Purge Gas:
               Purge Time:
               Purge Flow Rate:
               Purge Temperature:
Helium or Nitrogen
11.0 ±0.1 minute
25-40 mL/minute
Ambient temperature (required)
               Desorb Conditions
               Desorb Temperature:
               Desorb Flow Rate:
               Desorb Time:
180°C
15 mL/minute
4.0 ±0.1 minute
               Trap Reconditioning Conditions

               Reconditioning Temperature:
               Reconditioning Time:
180°C
7.0 ±0.1 minute (minimum).   A
longer time may be required to
bake-contamination or water
from the system.
9.1.1.2       Before initial use, condition the trap overnight at 180°C by
              backflushing with at least 20 milliliters (mL)/minute flow of
              inert gas.  Do not vent the trap effluent onto the analytical
              column.  Prior to daily use, condition the trap by heating at
              180°C for 10 minutes while backflushing.  The trap may be vented
              to the analytical column during daily conditioning; however, the
              column must be run through the temperature program prior to the
              analysis of samples and blanks.

9.1.1.3       Optimize purge and trap conditions for sensitivity and to minimize
              cross-contamination between samples.  Once optimized, the same
                                         D-21/VOA
                              OLC03.2

-------
Exhibit D Volatiles -- Section 9
Calibration and Standardization  (Con't)

              purge and trap conditions must be used for the analysis of all
              standards,  samples, and blanks.

9.1.1.4       A moisture reduction/water management system may be used to
              improve the chromatographic performance by controlling moisture or
              water if:

              •    The system does  not  introduce contaminants  which.interfere
                   with identification  and quantitation  of  compounds listed  in
                   Exhibit  C (Volatiles) ,

              •    The analytical results  generated when using the moisture
                   reduction/water  management  system meet the  initial and
                   continuing calibration technical acceptance criteria listed in
                   the Statement  of Work (SOW)  and  the Contract Required
                   Quantitation Limits  (CRQLs)  listed.in Exhibit C (Volatiles);

              •    All calibration  standards,  samples, and blanks are analyzed
                   under  the same conditions;  and

              •    The Contractor performs acceptably on the Performance
                   Evaluation Samples using  this system.

9.1.2      Gas Chromatograph (GC)

9.1.2.1       The  following  are  the recommended GC analytical conditions.  The
              conditions are recommended unless otherwise noted:

                Capillary Columns
                Carrier Gas:                    Helium
                Flow Rate:                      15  mL/minute
                Initial Temperature:            10°C
                Initial Hold Time:              1.0 - 5.0  (±0.1) minutes
                Ramp Rate:                      6°C/minute
                Final Temperature:              160°C
                Final Hold  Time:                Until three minutes after all
                                               compounds listed in Exhibit C
                                                (Volatiles) elute  (required)

9.1.2.2       Optimize GC  conditions for  analyte separation and sensitivity.
              Once optimized,  the  same GC conditions must be  used for  the
              analysis of  all  standards,  samples,.and blanks.

9.1.2.3       If the gaseous compounds chloromethane, bromomethane,  vinyl
              chloride, and chloroethane  fail  to exhibit narrow,  symmetrical
              peak shape,  are  not  separated from the  solvent  front,  or are not .
              resolved greater than 90.0  percent from each other,  then a
              subambient oven controller  must  be used,  and the  initial
              temperature  must be  less than or equal  to 10°C.
 OLC03.2                                 D-22/VOA

-------
                                                         Exhibit D Volatiles -- Section 9
                                                 •Calibration and Standardization  (Con't)

9.1.3     Mass Spectrometer  (MS)

          The following are the required MS analytical conditions:

           Electron  Energy:             70  volts (nominal)
           Mass Range:                  35-300 amu
           lonization Mode:             El
           Scan Time:                   To  give at  least  five scans per peak,  not
                                        to  exceed 2 seconds per scan for capillary
                                        column.


9.2    Instrument Performance Check -- 4-Bromofluorobenzene (BFB)

9.2.1     Summary of Instrument Performance Check

9.2.1.1       The GC/MS system must be tuned to meet the  manufacturer's
              specifications,  using a suitable calibrant  such as perfluoro-tri-
              n-butylamine (FC-43)  or perfluorokerosene (PFK).   The mass
              calibration and resolution of the GC/MS system are verified by the
              analysis of the instrument performance check solution (Section
              7.2.3.1) .

9.2.1.2       Prior to the analysis of any  samples, blanks, or calibration
              standards,  the Contractor must establish that the GC/MS system
              meets the mass spectral ion^ abundance criteria for the instrument
              performance check solution containing BFB.

9.2.2     Frequency  of Instrument  Performance Check           ,

          The instrument  performance  check solution must be  injected once at
          the beginning of  each 12-hour period,  during which samples,  blanks or
          standards  are to  be analyzed.  The  12-hour time  period  for GC/MS
          performance check,  calibration standards (initial  or continuing
          calibration), blank, and sample  analysis begins  at the  moment  of
          injection  of the  BFB analysis that  the laboratory  submits  as
          documentation of  a  compliant  instrument  performance  check.   The time
          period  ends after 12 hours  have  elapsed  according  to the  system
          clock.

9.2.3     Procedure  for  Instrument Performance Check

          The analysis of the instrument performance check solution may  be
          performed  as follows:

          •    As an injection of up to 50  nanograms  (ng) of BFB into the GC/MS
               (Section 7.2.3.1);  or

          •    By adding  50 ng of BFB to a  calibration standard (Section
               7.2.3.2) and analyzing the calibration standard.
                                         D-23/VOA                                 OLC03.2

-------
Exhibit D Volatiles -- Section 9
Calibration and Standardization  (Con't)

9.2.4     Technical Acceptance Criteria for Instrument Performance Check

9.2.4.1       The mass spectrum of BFB must be acquired in the following manner.
              Three scans (the peak apex scan and the scan immediately preceding
              and the scan immediately following the apex) are acquired and
              averaged.  Background subtraction is required,  and must be
              accomplished using a single scan no more than 20 scans prior to
              the beginning of the elution of BFB.  Do not background subtract
              part of the BFB peak.

              NOTE: All subsequent standards,  samples, and blanks associated
              with a BFB analysis must use identical MS instrument conditions.

9.2.4.2       The analysis of the instrument performance check solution must
              meet the ion abundance criteria given in Table D-l.

9.2.5     Corrective Action for  Instrument Performance Check

9.2.5.1       If the BFB technical acceptance criteria are not met, retune the
              GC/MS system.  It may also be necessary to clean the ion source,
              clean the quadrupole rods, or take other corrective actions to
              achieve the technical acceptance criteria.

9.2.5.2       BFB technical acceptance criteria must be met before any
              standards, samples, or required.blanks are analyzed.  Any samples
              or required blanks analyzed when tuning technical acceptance
              criteria have not been met will require re-analysis at no
              additional cost to USEPA.

9.3    Initial Calibration

9.3.1      Summary of Initial Calibration

           Prior to the analysis of samples and required blanks and after the
           instrument performance check technical acceptance criteria have been
           met,  each GC/MS system must be calibrated at a minimum of five '
           concentrations to determine instrument sensitivity and the linearity
           of GC/MS response for the purgeable target and Deuterated Monitoring
           Compounds  (DMCs).

9.3.2      Frequency of Initial Calibration

9.3.2.1       Each GC/MS system  must be calibrated upon award of the contract,
              whenever the  Contractor  takes corrective action which may change
              or affect  the  initial calibration  criteria  (i.e.,  ion source
              cleaning or repair,  column  replacement,  etc.),  or  if the
              continuing calibration  technical acceptance criteria have not  been
              met.

 9.3.2.2       If  time remains  in the  12-hour  time period  after meeting the
              technical  acceptance criteria  for  the  initial  calibration,  samples
              and  blanks may be  analyzed.   It is .not necessary to  analyze a
              continuing calibration  standard if the initial calibration
              standard that is the same concentration as  the continuing
             ' calibration'standard meets  the  continuing calibration technical

OLC03.2                                   D-24/VOA

-------
                                                         exhibit D Volatiles -- Section 9
                                                  Calibration and Standardization (Con't)

              acceptance  criteria.   A method blank is  required.   Quantitate all
              samples and blank results  against the initial  calibration standard
              that  is the same  concentration as the continuing calibration
              standard.   Compare Quality Control (QC)  criteria such as internal
              standard area response change and retention time shift to the
              initial calibration standard that is the same  concentration as the
              continuing  calibration standard.

9.3.3     Procedure  for  Initial Calibration

9.3.3.1       Assemble a  purge  and trap  device that meets the specifications in
              Section 6.4.  Condition the device as described in Section 9.1.1.

9.3.3.2       Connect the purge and trap device to the GC.   The GC must be
              operated using temperature and flow rate parameters equivalent to
              those in Section  9.1.2.

9.3.3.3       All samples, blanks,  and standard/spiking solutions must be
              allowed to  warm to ambient temperature (approximately 1 hour)
              before analysis.

9.3.3.4       Add 10 microliters (pL)  of the internal  standard solution (Section
              7.2.3.3) to each  aqueous standard containing the DMCs for a
              concentration of  5 micrograms per liter  (ug/L)  at time of purge.
              Analyze each calibration standard according to Section 10.2.

9.3.4     Calculations for Initial  Calibration

          Calculating the Relative  Response  Factor (RRF) of  the xylenes
          requires  special attention.   On capillary columns,  the — and p-xylene
          isomers coelute.   Therefore,  when  calculating  the  relative  response
          factor in the  equation below,  use  the area  response (Ax)  and         .          >
          concentration  (Cx) of the peak from  o-xylene.

9.3.4.1       Calculate RRF for each purgeable target  compound and DMC using
              Equation 1.  See  Table D-3 to associate  purgeable target compounds
              and DMCs with the proper internal standard.  See Table D-4 for
              primary quantitation ions  to be used for each purgeable target
              compound, DMC, and internal standard compound.

              NOTE:  Unless otherwise stated the area response is that of the
              primary quantitation ion.
                                         D-25/VOA                                 OLC03.2

-------
Exhibit D Volatiles -- Section 9
Calibration and Standardization (Con't)

              EQ.  1

                                     MT- _^L« fii
              Where:

               AX   =   Area of the characteristic ion (EICP)  for the compound to
                       be measured (Table D-4).

               Ais   =   Area of the characteristic ion (EICP)  for the specific
                       internal standard (Tables D-3 and D-4).

               Cis   =   Concentration of the internal standard.

               Cx   =   Concentration of the compound to be measured.
                                               • • •
9.3.4.2       The mean Relative Response Factor  (RRF) must be calculated for all
              compounds.

9.3.4.3       Calculate the Percent Relative Standard Deviation  (%RSD)  of RRF
              values for each purgeable target and DMC over the initial
              calibration range using Equation 2 in conjunction with Equation 3.
              EQ. 2
                                            SD__
                                             J?* » 100
                                             x
              Where:

               SDRRF  =   Standard deviation of initial calibration relative
                         response factors (per compound).   From EQ.  3.

               •      =   Mean value of the initial calibration relative response
                         factors (per compound).

 9.3.4.4       Equation 3 is the general formula for Standard Deviation  (SD) for
              a statistically small set of values.

              EQ. 3
                                              n - 1
              Where:

                Xi    =  Each individual value used to calculate the mean.

                •     =  The mean of n values.

                n     =  Total number of values.


 OLC03.2                                 D-26/VOA

-------
                                                         Exhibit D Volatiles -- Section 9
                                                  Calibration and Standardization  (Con't)
9.3.4.5       Equation 4 is the general formula for the mean of a set of values.

               EQ.  4

                                            *^
                                                X.
               Where:
                xi    =  Value.

                •    =  Mean value.

                n    =  Number of values.

9..3.5     Technical Acceptance Criteria For  Initial Calibration

9.3.5.1       All initial calibration standards must be analyzed at the
              concentration levels described in Section 7.2.3.6.1, and at the
              frequency described in Section 9.3.2 on a GC/MS system meeting the
              BFB technical acceptance criteria (Section 9.2.4).

9.3.5.2       Excluding those ions in the solvent front, no quantitation ion may
              saturate the detector.  Consult the instrument manufacturer's
              instrument operating manual to determine how saturation is
              indicated for your instrument.

9.3.5.3       The RRF at each calibration concentration for each purgeable
              target and DMC that has a required minimum relative response
              factor value must be greater than or equal to the compound's
              minimum acceptable relative response factor listed in Table D-2.

9.3.5.4-      The %RSD for each target or DMC listed in Table D-2 must be less
              than or equal to that value listed.

9.3.5.5       Up to two compounds may fail the criteria listed in Sections
              9.3.5.3 and 9.3.5.4 and still meet the minimum RRF and %RSD
              requirements.  However, these compounds must have a minimum RRF
              greater than or equal to 0.010, and the %RSD must be less than or
              equal to 40.0 percent.

9.3.6     Corrective  Action for  Initial Calibration

9.3.6.1       If the initial calibration technical acceptance, criteria are not
              met,  inspect the system for problems.  It may be necessary to
              clean the ion source, change the column, service the purge and
              trap device, or take other corrective actions to achieve the
              technical acceptance criteria.

9.3.6.2       Initial calibration technical acceptance criteria MUST be met
              before any samples or required blanks are analyzed.  Any samples
                                         D-27/VOA     .                            OLC03.2

-------
Exhiibit D Volatiles -- Section 9
Calibration and Standardization  (Con't)

              or required blanks analyzed when initial calibration technical
              acceptance criteria have not been met will require re-analysis at
              no additional cost to USEPA..

9.4    Continuing Calibration

9.4.1     Summary  of  Continuing Calibration

          Prior to the  analysis of samples and required blanks and after  BFB
          and initial calibration technical  acceptance criteria have  been met,
          each GC/MS  system must be routinely checked by analyzing a  continuing
          calibration standard containing all the  purgeable target and DMCs and
          internal standards to ensure that  the instrument continues  to meet
          the instrument sensitivity and linearity requirements of the SOW.

9.4.2     Frequency of  Continuing Calibration

9.4.2.1       A check of the calibration curve must be performed once per every
              12-hour time period of operation.  The 12-hour time period begins
              with the injection of BFB.

9.4.2.2       If time remains in the 12-hour time period after meeting the
              technical acceptance criteria for the initial calibration, samples
              and blanks may be analyzed.  It is not necessary to analyze a
              continuing calibration standard,  if the  initial calibration
              standard that is  the same  concentration  as the continuing
              calibration  standard meets the continuing calibration technical
              acceptance criteria.  A method blank  is  required.  Quantitate all
              sample  and blank  results against  the  initial  calibration standard
              that  is the  same  concentration as the continuing  calibration
              standard  (5.0 ug/L  for non-ketones, 25 ug/L  for ketones).

              If time does not  remain  in the  12-hour period beginning with  the
              injection of the  instrument performance  check solution, a new
              injection of the  instrument performance  check solution must be
              made.   If the new injection meets the ion abundance  criteria  for
              BFB,  then a  continuing calibration standard  may be injected.

9.4.3      Procedure for Continuing Calibration

9.4.3.1       Set  up  the purge  and trap  GC/MS  system per the requirements in
              Section 9.1.

9.4.3.2.      All  samples, required blanks,  and standard/spiking solutions  must
              be allowed to  warm to ambient  temperature  (approximately 1 hour)
              before  analysis.

9.4.3.3       Add  10  uL of internal  standard spiking  solution  (prepared as
               described in Section 7.2.3:3 to the  25  mL syringe or volumetric
               flask containing the continuing calibration  standard in Section
               7.2.3.6.3).  Analyze the continuing  calibration  standard,
               according to Section 10.2.
 OLC03.2                                 D-28/VOA

-------
                                                         Exhibit D Voiatiles -- Section 9
                                                  Calibration and Standardization (Con't)

9.4.4     Calculations for Continuing Calibration

9.4.4.1       Calculate a RRF for each target  and DMC according to Section
             ' 9:3~. 4.1.'  - ' 	   ~                   -   ~

9.4.4.2       Calculate the  percent  difference between the continuing
              calibration RRF and the most recent initial  calibration mean RRF
              for each purgeable target and DMC using Equation 5.

              EQ. 5

                                            RRT  -  1ESP7
                            IDilteeeaee •  	'       * - 100
              Where:

              RRFC    =    Relative response factor from current continuing
                           calibration standard.

              	          Mean relative response factor from the most recent
             • 1?RFi   =     initial calibration.

9.4.5     Technical Acceptance Criteria for Continuing Calibration

9.4.5.1       The concentration of the volatile organic  target and deuterated
              monitoring compounds in the continuing calibration standard must
              be 5.0  ug/L for non-ketones and 25  ug/L for ketones.   The
              continuing calibration standard must be analyzed at the frequency
              described in Section 9.4.2 on a GC/MS system meeting the BFB
              (Section 9.2.4)  and the initial calibration (Section 9.3.5)
              technical acceptance criteria.

9.4.5.2       The RRF for each purgeable target and DMC  that has a required
              minimum RRF value must be greater than or  equal to the compound's
              minimum acceptable RRF listed in Table D-2.

9.4.5.3       The RRF percent difference for each purgeable target and DMC
              listed  in Table D-2 must be less than or equal to that value
              listed.

9.4.5.4       Up to two compounds may fail the requirements listed in Sections
              9.4.5.2 and 9.4.5.3 and still meet  the minimum RRF criteria and
              percent difference criteria.  However,  these compounds must have  a
              minimum RRF greater than or equal to 0.010  and the percent
              difference must be within the inclusive range of ±40.0 percent.

9.4.5.5       Excluding those ions in the solvent front,  no quantitation ion may
              saturate the detector.  Consult the manufacturer's instrument
              operating manual to determine how saturation is indicated for your
              instrument.
                                         D-29/VOA                                  OLC03.2

-------
Exhibit D Volatiles -- Section 9
Calibration and Standardization  (Con't)

9.4.6     Corrective Action  for  Continuing Calibration

9.4.6.1       If the continuing calibration technical acceptance criteria are
              not met,  recalibrate the GC/MS instrument according to Section
              9.3.  It may be necessary to clean the ion source, change the
              column,  or take other corrective actions to achieve the continuing
              calibration technical acceptance criteria.

9.4.6.2       Continuing calibration technical acceptance criteria MUST be met
              before any samples or required blanks are analyzed.  Any samples
              or required blanks analyzed when continuing calibration technical
              acceptance criteria have not been met will require re-analysis at
              no additional cost to USEPA.
 OLC03.2                                 D-30/VOA

-------
                                                        Exhibit D Volatiies -- Section 10
                                                                          Sample Analysis

10.0   SAMPLE ANALYSIS

10.1   Summary of Sample Analysis

10.1.1    This method is designed  for  analysis of samples that contain low
          concentrations of  the  target compounds listed in Exhibit C.  It is
          expected that the  samples will come from drinking water and
          well/ground water  type sources around Superfund sites.  If, upon
          inspection of a  sample,  the  Contractor suspects that the sample is
          not amenable to  this method,  contact Sample Management Office  (SMO).
          SMO will contact the Region  for  instructions.

10.1.2    Prior  to the analysis  of samples, establish the appropriate purge and
          trap Gas Chromatograph/Mass  Spectrometer  (GC/MS) operating
          conditions, as outlined  in Section 9.1, analyze the instrument
          performance check  solution  (Section 9.2), and calibrate the GC/MS
          system according to Sections 9.3  through 9.4.6.  Also prior to sample
          analysis, a method blank must be  analyzed that meets blank technical
          acceptance criteria in Section 12.1.5.  All samples, required blanks,
          and standard/spiking solutions must be allowed to warm to ambient
          temperature  (approximately 1 hour) before analysis.  All samples,
          required blanks, and calibration  standards must be analyzed under the
          same instrument  conditions.

10.1.3    If insufficient  sample volume (less than 90 percent of the required
          amount) is received to perfprm the analyses, the Contractor shall
          contact SMO to apprise them  of the problem.  SMO will contact the
          Region for instructions. The Region will either require that no
          sample analyses  be performed or will require a reduced volume be used
          for the sample analysis,. No other changes in the analyses will be
          permitted.  The  Contractor shall  document the Region's decision in
          the Sample Delivery Group  (SDG) Narrative.

10.2   Procedure for Sample Analysis

10.2.1    Remove the plunger from  a 25 milliliters  (mL) syringe that has a
          closed syringe valve attached.  Open the sample or standard container
          which  has been allowed to come to ambient temperature, and carefully
          pour the sample  into the syringe  barrel to just short of overflowing.
          Replace the syringe plunger  and compress the sample.  Invert the
          syringe, open the  syringe valve,  and vent any residual air while
          adjusting the sample volume  to 25.0 mL.  This process of taking an
          aliquot destroys the validity of  the sample for future analysis,
          unless the excess  sample is  immediately transferred to a smaller vial
          with zero headspace and  stored at 4°C  (±2°C).

10.2.2    Once the sample  aliquots have been taken from the VOA vial, the pH  of
          the water sample must  be determined.  The purpose of the pH
          determination is to ensure that all VOA samples were acidified in the
          field.  Test the pH by placing one or two drops of sample on the pH
          paper  (do not add  pH paper to the vial).  Record the pH of each
          sample and report  these  data in the SDG Narrative, following the
          instructions in  Exhibit  B.   No pH adjustment is to be performed by
          the Contractor.
                                         D-31/VOA
OLC03.2

-------
Exhibit D Volatiles -- Section 10
Sample Analysis  (Con't)

          Add  10.0 microliters  (uL) of the  internal  standard  spiking solution
          and  10.0 uL  of  the  Deuterated Monitoring Compound  (DMC)  standard
          solution through the valve bore of  the  syringe,  then  close the valve.
          Invert the syringe  three times.

10.2.3    Attach the valve assembly on the  syringe to the  valve on the  sample
          purger.  Open the valves and inject the sample into the  purging
          chamber.

10.2.4    Close both valves and  purge the sample  for 11.0  (±0.1) minutes at
          ambient temperature.

10.2.5    Sample Desorption - After the 11-minute purge,  attach the trap to
          the  GC, adjust  the  purge and trap system to the  desorb mode,  initiate
          the  temperature program sequence  of the GC and start  data
          acquisition.  Introduce the trapped material to  the GC column by
          rapidly heating the trap to 180°C while backflushing  the trap with
          inert gas  at 15 mL/min for 4.0 ±0.1 min.   While  the trapped material
          is being  introduced into the GC,  empty  the sample  purger and  rinse it
          with reagent water. For samples containing large amounts of water-
          soluble  materials, suspended  solids, high boiling compounds, or  high
          purgeable levels, it may be necessary tp wash out  the sample  purger
          with a detergent solution, rinse  it with reagent water,  and then  dry
          it in an  oven at 105°C.

10.2.6    Trap Reconditioning -   After  desorbing  the sample,  recondition  the
          trap for  a minimum of 7.0  ±0.1 min at 180°C by returning the  purge
          and trap  system to purge mode.

10.2.7    Gas  Chromatograohv - Hold  the  column temperature at 10°C for  1.0  to
           5.0  min,  then program at  6°C/min to 160°C  and hold until three
          minutes after all target volatile compounds have eluted.

          NOTE: Once an initial hold time  has been chosen and the  GC operating
           conditions optimized,  the  same GC condition must be used for  the
           analysis.

10.2.8     Termination of Data Acquisition -  Three minutes after all the
           purgeable target compounds have  eluted from the GC, terminate the MS
           data acquisition and store data files on the data  system storage
           device.  Use appropriate  data output software to display full range
           mass spectra and appropriate Extracted Ion Current Profiles (EICPs).

10.2.9     Dilutions

10.2.9.1      An original undiluted analysis  must be made  and results reported
               for all  samples.   If the on-column  concentration  of  any target
               compound in any sample exceeds  the  initial calibration range,  a
               new aliquot of  that sample must be  diluted and  purged.  Guidance
               for performing  dilutions and  exceptions to this requirement are
               given  in Sections  10.2.9.2 through  10.2.9.8.

               NOTE  1:  If  the  laboratory has evidence or highly  suspects, because
               of sample  color or other physical property,  that  a sample may
               contain  high concentrations of  either  target or non-target

OLC03.2                                 D-32/VOA

-------
                                                        Exhibit D Volatiles  -- Section  10
                                                                  Sample Analysis  (Con't)

              compounds, then SMO shall be contacted immediately.   SMO will seek
              regional recommendations for diluted analysis.

              NOTE 2: Secondary ion quantitation is only allowed when there are
              sample interferences with the primary quantitation ion, not when
              saturation occurs.  If secondary ion quantitation is used,
              calculate a Relative Response Factor (RRF) using the area response
              (EICP) from the most intense secondary ion which is free of sample
              interferences,  and document the reasons in the SDG Narrative.

10.2.9.2      Use the results of the original analysis to determine the
              approximate dilution factor required to get the largest analyte
              peak within the initial calibration range.

10.2.9.3      The dilution factor chosen should keep the concentration of the
              volatile target compounds that required dilution in the upper half
              of the initial calibration range.

10.2.9.4      All dilutions must be made just prior to GC/MS analysis of the
              sample.  Until the diluted sample is in a gas-tight syringe,  all
              steps in the dilution procedure must be performed without delay.

10.2.9.5      Samples may be diluted in a volumetric flask or in a 25 mL
              "Luerlock" syringe.

..10.2.9.6      To dilute the sample in a volumetric flask, use the following
              procedure:
•r i .
10.2.9.6.1        Select  the volumetric flask that  will  allow for necessary
'-"•                dilution (25  mL to 100 mL) .

:l'0.2.9.6.2        Calculate the approximate volume  of  reagent water which will  be
                  added to the  volumetric flask selected and add slightly less
                  than this quantity of reagent water  to the flask.

10.2.9.6.3        Inject  the proper sample aliquot  from  a syringe into the
                  volumetric flask.   Only aliquots  of  1  mL increments  are
                  permitted.  Dilute the aliquot to the  mark with reagent water.
                  Cap the flask and invert it three times.

10.2.9.6.4        Fill a  25 mL  syringe with the diluted  sample and analyze
                  according to  Section 10.2.

10.2.9.7      To dilute the sample  in a  25 mL  syringe, use the following
              procedure:

10.2.9.7.1        Calculate the volume of the reagent  water necessary  for the
                  dilution.  The final volume of the diluted sample should be 25
                  mL.

10.2.9.7.2        Close the syringe valve,  remove the  plunger from the syringe
                  barrel,  and pour reagent water into  the syringe barrel  to just
                  short of overflowing.
                                         D-33/VOA                                  OLC03.2

-------
 Exhibit D Volatiles -- Section 10
 Sample Analysis (Con't)

'10.2.9.7.3       Replace the syringe plunger and compress the water.

 10.2.9.7.4       Invert the syringe, open the syringe valve,  and vent any
                  residual air.  Adjust the water volume to the desired amount.

 10.2.9.7.5       Adjust the plunger to the 25 tnL mark to accommodate the sample
                  aliquot.  Inject the proper aliquot of sample from another
                  syringe through the valve bore of the 25 mL syringe.  Close the
                  valve and invert three times.  Analyze according to Section
                  10.2.

 10.2.9.8      For  total  xylenes where three isomers are quantified as two peaks,
               the  calibration of each peak should be considered separately,
               e.g., a diluted analysis is not required for total xylenes unless
               the  concentration of the peak representing the  single isomer
               exceeds 25 micrograms per liter (ug/L)  on-column,  or the peak
               representing the two co-eluting isomers exceeds 50 ug/L on-column.
 OLC03.2                                  D-34/VOA

-------
                                                        Exhibit D Volatiles -- Section 11
                                                           Data Analysis and Calculations
11.0   DATA ANALYSIS AND CALCULATIONS

11.1   Qualitative Identification of Target Compounds

11.1.1    The compounds listed  in  the Target Compound List  (TCL), Exhibit  C
           (Volatiles), shall be identified by an analyst competent  in the
          interpretation of mass spectra by comparison of the  sample mass
          spectrum  to the mass  spectrum.of a standard of the suspected
          compound.   Two criteria  must be satisfied  to verify  the
          identifications:
           •    Elution of the sample component at the same Gas Chromatograph
               (GO Relative Retention Time  (RRT) as the standard component; and

           •    Correspondence of the sample component and calibration standard
               component mass spectra.

•11.1.2     For establishing  correspondence of the GC RRT,  the sample component
           RRT must be within ±0.06 RRT units of  the RRT of  the standard
           component.   For reference,  the standard must  be run in the same 12-
           hour time period  as the  sample.  If samples are analyzed during the
           12-hour time period as the  initial calibration, use the RRT values
           from the 5 micrograms per liter (ug/L)  standard.   If co-elution of
           interfering compounds prohibits accurate  assignment of the sample
           component RRT  from the total ion chromatogram,  then the RRT should be
           assigned using the Extracted Ion Current  Profile  (EICP)  for ions
           unique to the  component  of  interest.

11.1.3     For comparison of standard  and sample  component mass spectra,  mass
    \     spectra obtained  on the  Contractor's GC/MS are required.  Once
           obtained, these standard spectra may be used  for  identification
           purposes, only if the Contractor's GC/MS  meets the daily instrument
           performance requirements for 4-Bromofluorobenzene (BFB).   These
           standard spectra  may be  obtained from  the standard analysis that was
           also used to obtain the  RRTs.

11.1.4     The guidelines for qualitative verification by comparison of  mass
           spectra are as follows:

11,. 1.4.1      All  ions present  in  the standard mass spectra at a relative
              intensity greater than 10 percent  (most abundant ion in the
              spectrum equals 100  percent) must be present in the sample
              spectrum.

11.1.4.2      The  relative  intensities of  ions specified in Section 11.1.4.1
              must  agree within ±20 percent between the standard and sample
              spectra.   (Example:  For an  ion with an abundance of 50 percent in
              the  standard  spectra, the corresponding sample abundance must be
              between 30 and 70 percent.)

11.1.4.3      Ions  greater  than 10 percent in the sample spectrum but not
              present in the standard spectrum must be considered and accounted
              for  by the analyst making the comparison.   The verification
              process should favor false positives.   All compounds meeting the
              identification criteria must be reported with their spectra.  For

                                         D-35/VOA
                                                                                  OLC03.2

-------
Exhibit D Volatiles -- Section 11
Data Analysis and Calculations  (Con't)

              all compounds below the Contract Required Quantitation Limit
              (CRQL),  report the actual value followed by a "J" (e.g., "0.3J").

11.1.4.4      If a compound cannot be verified by all of the spectral
              identification criteria listed in Section 11.1.4, but in the
              technical judgment of the mass spectral interpretation specialist,
              the identification is correct, then the Contractor shall report
              that identification and proceed with quantitation.

11.2   Qualitative Identification of Non-Target Compounds

11.2.1     A library  search shall be executed for non-target sample components
           for the purpose of tentative identification.   The NIST/EPA/NIH (May
           1992  release or later)  and/or Wiley (1991 release or later)  or
           equivalent  mass spectral library, shall be used as the  reference
           library.

11.2.2     Up to 30 organic compounds of greatest apparent concentration not
           listed in  Exhibit C for the volatile or semivolatile organic
           fraction,  excluding the Deuterated Monitoring Compounds (DMCs) and
           internal standard compounds, shall be tentatively identified via a
           forward search of the NIST/EPA/NIH (May 1992 release or later) and/or
           Wiley (1991 release or later), or equivalent mass spectral library.
           The following are not to be reported:

           •  Compounds with a response of  less than  10 percent of the internal
              standard (as determined  by inspection of the peak areas or
              heights);

           •  Compounds which elute  earlier than  30 seconds before the first
              purgeable compound listed in  Exhibit C  (Volatiles) or  three
              minutes  after  the  last purgeable  compound  listed  in Exhibit  C
               (Volatiles)  are not  required  to be  searched in  this fashion;

           •  Carbon dioxide; and

           •  Semivolatile TCL  compounds listed in Exhibit  C.

           Only after visual comparison of sample spectra with the nearest
           library searches will the mass spectral interpretation specialist
           assign a tentative identification.

           NOTE: Computer-generated  library search routines must not use
           normalizations which would misrepresent the library or unknown
           spectra when compared to  each other.

 11.2.3    Up to 20 peaks of greatest apparent concentration  (as determined by
           inspection  of peak areas  or  heights) that are suspected to be
           straight-chain, branched, or cyclic alkanes, alone or part of an
           alkane series,  shall be library searched.  Documentation for  the
           tentative identification  must be supplied.  Alkanes are not counted
           as part of  the  30  organic compounds described in Section 11.2.2.
 OLC03.2                                  D-36/VOA

-------
                                                        Exhibit D Volatiles -- Section 11
                                                   Data Analysis and Calculations (Con't)

11.2.4    Guidelines for making tentative identification:

11.2.4.1      All  major ions present in the  reference mass spectrum at  a
              relative  intensity greater than 10 percent' "(most 'abundant'ion in
              the  spectrum equals  100 percent)  must be present in the sample
              spectrum.

11.2.4.2      The  relative intensities of the major ions specified in Section
              11.2.4.1  must agree  within ±20 percent between the reference and
              sample spectra.  (Example: For an ion with an abundance of 50
              percent in the reference spectrum, the corresponding sample ion
              abundance must be between 30 and 70 percent.).

11.2.4.3      Molecular ions present in the  reference spectrum should be present
              in the sample spectrum.

11.2.4.4      Ions present in the  sample spectrum but not in the reference
              spectrum should be reviewed for possible background contamination
              or presence of coeluting compounds.

11.2.4.5      Ions present in the  reference  spectrum but not in the sample
              spectrum should be reviewed for possible subtraction from the
   ',;•-•         sample spectrum because of background contamination or coeluting
              compounds.  Data system library reduction programs can sometimes
              create these discrepancies.

11.2.4.6      Non-target compounds receiving a library search match of  85
              percent or higher should be considered a "probable match".  The
              compound should be reported unless the mass spectral
              interpretation specialist feels there is just evidence not to
              report the compound  as identified by the library search program.
   ~i~         The  lab should include in the  Sample Delivery Group (SDG)
              Narrative the justification for not reporting a compound  as listed
              by the search program.

11.2.4.7      If the library search produces more than one compound at  or above
              85 percent, report the compound with the highest percent  match
              (report first compound if percent match is the same for two or
              more compounds), unless the mass spectral interpretation
              specialist feels that the highest match compound should not be
              reported, or another compound with a lower match should be
              reported.  The lab should include in the SDG Narrative the
              justification for not reporting the compound with the highest
              spectral match.

11.2.4.8      If the library search produces a series of obvious isomer
              compounds with library search matches greater than 85 percent
              (e.g., tetramethyl naphthalenes), the compound with the highest
              library search percent match should be reported (or first compound
              if library search matches are the same).  A note should be placed
              in the SDG Narrative indicating the exact isomer configuration,  as
              reported, may not be accurate.

11.2.4.9      If the library search produces no matches at or above 85  percent
              and  in the technical judgement of the mass spectral interpretation

                                         D-37/VOA                                 OLC03.2

-------
Exhibit D Volatiles -- Section 11
Data Analysis and Calculations (Con't)

              specialist,  no valid tentative identification can be made,  the
              compound should be reported as unknown.   The mass spectral
              specialist should give additional classification of the unknown
              compound,  if possible (e.g., unknown aromatic, unknown
              hydrocarbon, unknown acid type, unknown chlorinated compound).  If
              probable molecular weights can be distinguished, include them.

11.2.4.10     Straight-chain, branched, or cyclic alkanes are not to be reported
              as tentatively identified compounds on Form I LCV-TIC.  When the
              above alkanes are tentatively identified, the concentration(s) are
              to be estimated as described in Section 11.3.2 and reported in the
              SPG Narrative as alkanes. bv class (i.e., straight-chain,
              branched,  or cyclic, as a series, as applicable).

11.3   Calculations

11.3.1     Target Compounds

11.3.1.1      Target compounds identified shall be quantified by the internal
              standard method using Equation 6.  The internal standard used
              shall be that which is assigned  in Table D-3.  The Relative
              Response Factor  (RRF) from the continuing calibration standard is
              used to calculate the concentration in the sample.  When a target
              compound concentration is below  its CRQL but the  spectra meets the
              identification criteria, report  the concentration with a "J".  For
              example, if the CRQL is  0.50 ug/L and a concentration of 0.30 ug/L
              is calculated, report as "0.30 J".  Report ALL  sample
              concentration data as UNCORRECTED for blanks.

              EQ. 6                                                    .

                                       in ug/L -

              Where:

               AX     =  Area of the characteristic ion (EICP) for the compound
                         to be measured.  The primary quantitation ions for the
                         target compounds, internal standards, and the DMCs are
                         listed in Table D-4.

               Ais    =  Area of the characteristic ion (EICP) for the internal
                         standard.  The target compounds are listed witli their
                         associated internal standards in Table D-3.

               Is     =  Amount of internal standard added in nanograms (ng).

               RRF   =  The relative response factor from the continuing
                         calibration standard.

               V0     =  Total volume of water purged, in milliliters  (mL).
 OLC03.2                                 D-38/VOA

-------
                                                        Exhibit D Volatiles -- Section 11
                                                   Data Analysis and Calculations  (Con't)


               Df    =  Dilution factor.  The dilution factor  for analysis of
                        water samples for volatiles by this method is defined
                        as  the ratio of the number of milliliters  (mL) of water
                        purged  (i.e., V0 above)  to the number of mL of the
                        original water sample used for purging.  For example,
                        if  5.0 mL of sample is diluted to  25.0 mL with reagent
                        water and purged, Df=25.0 mL/5 mL  = 5.0.  If no
                        dilution is performed, Df =1.0.

11.3.1.2      Xylenes (o-,  m-,  and p-  isomers)  are to be reported as  xylenes
              (total).   Because - and p-xylene isomers coelute on capillary
              columns,  special  attention must  be given to the quantitation of
              the xylenes.   The RRF determined in Section 9.4.4.1,  is based on
              the peak that represents the single isomer on the GC column (o-
              xylene on capillary columns).   In quantitating sample
              concentrations, use the areas on both peaks and the RRF.   The
              areas of the two peaks may be summed and the concentration
              determined,  or the concentration represented by each of the two
              peaks may be determined separately and then summed.

11.3.1.3      The stereoisomers,  trans-l,2-dichloroethene,  and cis-
      ^;       1,2-dichloroethene,  are to be reported separately.

11.3.1.4      The requirements listed in Sections 11.3.1.5 and 11.3.1.6 apply  to
              all standards, samples,  and blanks.

11.3.1.5      It is expected that situations will arise where the automated
              quantitation procedures in the GC/MS software provide
              inappropriate quantitation.   This normally occurs when there is
              compound co-elution,  baseline noise,  or matrix interferences.   In
              these circumstances the Contractor must perform a manual
              quantitation.  Manual quantitations are performed by integrating
              the area of the quantitation ion of the compound.  This
              integration shall only include the area attributable to the
              specific TCL, deuterated monitoring,  or internal standard
              compound.  The area integrated shall not include baseline
              background noise.   The area integrated shall not extend past the
              point where the sides of the peak intersect with the baseline
              noise.  Manual integration is not to be used solely to meet QC
              criteria, nor is it to be used as a substitute for corrective
              action on the chromatographic system.   Any instances of manual
              integration must be documented in the SDG Narrative.

11.3.1.6      In all instances where the data system report has been edited, or
              where manual integration or quantitation has been performed,  the
              GC/MS operator must identify such edits or manual procedures by
              initialing and dating the changes made to the report,  and shall
              include the integration scan range.  In addition, a hardcopy
              printout of the EICP of the quantitation ion displaying the manual
              integration shall be included in the raw data.   This applies to
              all compounds listed in Exhibit  C (Volatiles),  internal standard,
              and DMCs.
                                         D-39/VOA                                 OLC03.2

-------
Exhibit D Volatiles -- Section 11
Data Analysis and Calculations (Con't)

11.3.2    Non-Target  Compounds

11.3.2.1      An estimated concentration for non-target compounds tentatively
              identified shall be determined by the internal standard method.
              For quantitation,  the nearest internal standard free of
              interferences shall be used.

11.3.2.2      Equation 6 is also used for calculating non-target compound
              concentrations.  Total area counts (or peak heights)  from the
              total Reconstructed Ion Chromatograms (RICs)  are to be used for
              both the non-target compound to be measured (AX)  and the internal
              standard (Ais) .  A  RRF  of  1.0  is  to be assumed.  The value  from
              this quantitation shall be qualified by a "J" (estimate due to
              lack of a compound-specific relative response factor), and "N"
              (presumptive evidence of presence),  indicating the qualitative and
              quantitative uncertainties associated with this non-target
              compound. 'An estimated concentration must be calculated for all
              tentatively identified compounds as well as those identified as
              unknowns.

11.3.3     CRQL  Calculation

         .  Calculate the adjusted CRQL for volatiles by using Equation 7.

           EQ. 7

                             Adjusted   Comtrnct.   ve
                               CRQL        CKQU,     ~
           Where:

            Contract CRQL =     Exact CRQL values in Exhibit C of  the SOW.

                          V0  =   Total volume of water purged in milliliters.
                              •  NOTE: Must not exceed the contract sample volume.

                          Vc  =   Contract sample volume in milliliters (25 mL).

                          Df  =   Same as EQ.  6.
           NOTE: If the adjusted CRQL is less than the CRQL listed in Exhibit C
           (Volatiles), report the CRQL listed in Exhibit C (Volatiles).

 11.3.4    Deuterated Monitoring Compound Recoveries

 11.3.4.1      Calculate  the  concentration  of  each DMC using the same equation as
               used for target  compounds  (Equation 6).

 11.3.4.2      Calculate  the  recovery of  each  DMC in  all  samples and blanks using
               Equation 8.  Report  the recoveries on  appropriate forms.
 OLC03.2                                 D-40/VOA

-------
                                                        Exhibit D Volatiles -- Section 11
                                                   Data Analysis and Calculations (Con't)
              EQ.  8

                             Percent Recover? -  —- * 100
              Where:
                                                                  /•
               Qd   =   Concentration or amount determined by analysis.

               Qa   =   Concentration or amount added to  sample/blank.


11.3.5    Internal Standard Responses and Retention Times

          Internal standard responses and retention times in all samples and
          blanks must be evaluated during or immediately after data
          acquisition.  Compare the sample/blank internal standard responses
          and retention times to the continuing calibration internal standard
          responses and retention times.  For samples and blanks analyzed
          during the same 12-hour time period as the initial calibration
          standards, compare the internal standard responses and retention
          times against the 5 ug/L calibration standard.  The EICP of the
          internal standards must be monitored arid evaluated for each sample
       ,   and blank.

11.4   Technical Acceptance Criteria for Sample  Analysis

11.4.1    The sample must be analyzed on a GC/MS system meeting the BFB,
          initial calibration, continuing calibration,  and blank technical
          acceptance criteria.

11.4.2    The sample and any required dilution must be  analyzed within the
          contract holding time.

11.4.3    The sample must have an associated method blank meeting the blank
          technical acceptance criteria.

11.4.4    The percent  recovery of each of the DMCs in the sample must be within
          the acceptance windows in Table D-5.  Up to three DMCs per sample may
          fail to meet the recovery limits listed in Table D-5.

11.4.5    The EICP area for each of the internal standards in the sample must
          be within the inclusive range of ±40.0 percent of its response in the
          most recent  continuing calibration standard analysis.

11.4.6    The retention time shift for each of the internal standards in the
          sample must  be within ±0.33 minutes (20.0 seconds) of its retention
          time in the  most recent continuing calibration standard analysis.

11.4.7    The RRT of each of the DMCs in the sample must be within ±0.06 RRT
          units of its relative retention time in the most recent continuing •
          calibration  standard analysis.
                                         D-41/VOA                                 OLC03.2

-------
Exhibit D Volatiles'-- Section 11
Data Analysis and Calculations (Con't)

11.4.8    Excluding those ions in the solvent front, no ion may saturate the
          detector.  No target compound concentration may exceed the upper
          limit of the initial calibration range unless a more dilute aliquot
          of the  sample is  also  analyzed according to the procedures in Section
          10.2.9.

11.4.9    The Contractor must demonstrate that there is no carryover from a
          contaminated sample before  data from subsequent analyses may be
          submitted.  After a sample  that contains a target compound at a level
          exceeding the initial  calibration  range, or a non-target compound at
          a concentration greater than 100 ug/L, or saturated ions from a
          compound  (excluding the compound peaks in the solvent front), the
          Contractor must either:

11.4.9.1      Analyze an instrument blank immediately after the contaminated
              sample.   If an autosampler is used,  an instrument blank must also
              be analyzed using the same purge inlet that was used for the
              contaminated sample.  The. instrument blanks must meet the
              technical acceptance criteria for blank analysis (Section 12.1.5);
              or

11.4.9.2      Monitor the analyzed sample immediately after the contaminated
              sample for all the compounds that were in the contaminated sample
              and that exceeded the limits above.   The maximum carryover
              criteria are as follows: the sample must not contain a concen-
              tration above  the CRQL for the target compounds, or above 2 ug/L
              for the non-target compounds that exceeded the limits in the
              contaminated sample.  If an autosampler is used, the next sample
              analyzed using the same purge inlet that was used for the
              contaminated sample must also meet the maximum carryover criteria.

11.5    Corrective Action for Sample Analysis

11.5.1    Sample technical  acceptance criteria must be  met  before data are
           reported.   Samples contaminated from laboratory sources or any
           samples not meeting the sample technical acceptance criteria will
           require re-analysis at no additional cost to  USEPA.

11.5.2    Corrective actions for failure to meet  instrument performance checks,
           initial calibration,  continuing calibration,  and method blanks must
           be completed before the analysis of samples.

11.5.3    If the technical  acceptance criteria for any of the internal
           standards and DMCs are not met,  check calculations,  internal standard
           and DMC spiking solutions,  and instrument performance.   It may be
           necessary to bake-out the system to remove the water from the purge
           and trap transfer lines,  to recalibrate the instrument, or take other
           corrective action procedures to meet the technical acceptance
           criteria.

 11.5.4    Sample reruns performed as a result of suspected matrix interference
           beyond the scope  of the method will be evaluated on a case-by-case
           basis for payment purposes by the USEPA Contract Laboratory Program
           Project Officer  (CLP PO).  Send a copy of the SDG Narrative
 OLC03.2                                  D-42/VOA

-------
                                                        Exhibit D Volatiies -- Section 11
                                                   Data Analysis and Calculations  (Con't)

           (including your contract number), a description of the situation, and
          the requested action to the CLP  PO.

11.5.5    If the contractor needs to analyze more than one  (1) sample dilution
          other than the original analysis to have all the  target compounds
          within the initial calibration range, contact Sample Management
          Office (SMO).  SMO will contact  the Region for instruction.

11.5.6    All samples  to be reported to USEPA must meet the maximum carryover
          criteria in  Section 11.4.9.  If  any sample fails  to meet these
          criteria, each subsequent analysis must be checked for cross
          contamination.  The analytical system is considered contaminated
          until a sample has been analyzed that meets the maximum carryover
          criteria or  an instrument blank  has been analyzed that meets the
          technical acceptance criteria for blanks.
                                         D-43/VOA                                 OLC03.2

-------
Exhibit D Volatiles -- Section 12
Quality Control

12.0   QUALITY CONTROL (QC)

12.1   Blank Analyses

12.1.1    Summary  of  Blank Analyses

          There  are three  different  types of blanks required by  this  method.

12.1.1.1      Method Blank - 25 milliliters (mL)  of reagent water spiked with
              10.0 microliters (uL) internal standard solution and 10.0 uL
              Deuterated Monitoring Compound (DMC)  solution, and carried through
              the entire analytical procedure.   The purpose of the method blank
              is to determine the levels of  contamination associated with
              processing and analysis of samples.

12.1.1.2      Storage Blank - Upon receipt of the first samples in a Sample
              Delivery Group  (SDG), two 40 mL screw cap VOA vials with a
              PTFE-faced silicons  septum are filled with reagent water (80 mL
              total).  The vials are stored with the samples in the SDG under
              the  same conditions.  A 25.0 mL aliquot of this reagent water is
              spiked with a 10.0 uL internal standard solution and 10.0 uL of
              DMC  solution and analyzed after all samples in the SDG have been
              analyzed.  The  storage blank indicates whether contamination may
              have occurred during storage of samples.

12.1.1.3      Instrument Blank - 25 mL of reagent water spiked with 10.0 uL of
              internal standard solution and 10.0 uL of DMC solution and carried
              through the entire analytical procedure.  Instrument blanks are
              analyzed after  a sample/dilution which contains a target compound
              at a concentration greater than 25 micrograms per liter  (ug/L)
               (ketones 125 ug/L),  or a non-target compound at a concentration
              greater than 100 ug/L or saturated ions from a compound  (excluding
              the  compound peaks in the solvent front).  The results  from
             . instrument blank analysis indicate whether there is contamination
              from a previous sample.

12.1.2    Frequency of Blank Analyses

12.1.2.1      The  method blank must be analyzed at least once during  every 12-
              hour time period on  each Gas Chromatograph/Mass Spectrometer
             .  (GC/MS) system  used  for volatile analysis  (see Section  9.2.2 for
              the  definition  of the  12-hour time period).

12.1.2.2      The  method blank must be analyzed after the continuing  calibration
              standard and before  any samples or storage blanks are analyzed.
              The  method blank must be analyzed after the initial calibration
              sequence if  samples  are analyzed before the 12-hour time period
              expires.  A  method blank must be analyzed  in  each  12-hour time
              period  in which samples  (including dilutions) and  storage blanks
              from an SDG  are analyzed.

 12.1.2.3      A  minimum of one storage blank must be analyzed per SDG, after all
               samples for  the SDG  have been analyzed, unless the SDG  contains
              only ampulated  Performance Evaluation  (PE) samples.  Analysis of a
 OLC03.2                                 D-44/VOA

-------
                                                        Exhibit D Voiatiies -- Section 12
                                                                  Quality Control  (Con't)

              storage blank is not required for SDGs that contain only ampulated
              PE samples.

12.1.2.4      The Contractor must demonstrate that there is no carryover from
              contaminated samples before data from subsequent analyses may be
              used.   Samples may contain target compounds at levels exceeding
              the initial  calibration range or non-target compounds at
              concentrations greater than 100 ug/L, or ions from a compound that
              saturate the detector (excluding the compound peaks in the solvent
              front).  An instrument blank must be analyzed immediately after
              the contaminated sample (also in the same purge inlet if an
              autosampler is used)', or a sample that meets the maximum carryover
              criteria in Section 11.4.9 must be analyzed.  For these purposes,
              if the instrument blank meets the technical acceptance criteria
              for blank analysis or the sample meets the maximum carryover
              criteria, the system is considered to be uncontaminated.  If the
              instrument blank or sample does not meet the criteria (i.e.,
              contaminated), the system must be decontaminated.  Until an
              instrument blank meets the blank technical acceptance criteria or
              a sample meets the maximum carryover criteria, any samples
              analyzed since the original contaminated sample will require re-
              analysis at no additional expense to USEPA.

              NOTE:  Only the instrument blank which demonstrates that there was
              no carryover from the previous sample or the instrument blank that
              demonstrates that the system is clean (Section 12.1.5.7) must be
       •?.•:•     reported.  Instrument blanks analyzed during the instrument
       >-"     decontamination process which exceed the requirements listed in
              Section 11.4.9 do not need to be reported.
       T
12.1.3 '%•'  Procedure  for Blank Analyses

12.1.3.1      Spike 25 mL of reagent water with 10.0 uL of the internal standard
              solution  (Section 7.2.3.3), and 10.0 uL of the DMC solution
              (Section 7.2.3.4).

12.1.3.2      Prepare and analyze the blanks as described in Section 10.2.

12.1.4     Calculations for Blank Analyses

           Perform data analysis and calculations  according to Section 11.

12.1.5     Technical  Acceptance Criteria for Blank Analyses

12.1.5.1      All blanks must be analyzed on a GC/MS system meeting the 4-Bromo-
              fluorobenzene  (BFB), initial calibration, and continuing
              calibration technical acceptance criteria, and at the frequency
              described in Section 12.1.2.

12.1.5.2      The storage blank must be analyzed on a GC/MS system that also
              meets the technical acceptance criteria for the method blank.

12.1.5.3      The percent recovery of each of the DMCs in the blank must be
              within the acceptance windows in Table D-5.
                                         D-45/VOA                                 OLC03.2

-------
Exhibit D Volatiles -- Section 12
Quality Control (Con't)

12.1.5.4      The  EICP area for each of the internal  standards in the blank must
              be within the inclusive range of ±40.0  percent of its response in
              the  most recent continuing calibration  standard analysis.

12.1.5.5      The  retention time shift.for each of the internal standards in the
              blank must be within ±0.33 minutes (20.0 seconds) of its retention
              time in the most recent continuing calibration standard analysis.

12.1.5.6      The  Relative Retention Time (RRT) of each of the DMCs in the blank
              must be within ±0.06 RRT units of its relative retention time in
              the  most recent continuing calibration standard analysis.

12.1.5.7      The  concentration of each target compound found in the storage and
              method blanks must be less than its CRQL listed in Exhibit C
              (Volatiles), except for methylene chloride and cyclohexane which
              must be less than 10 times their respective CRQLs, and acetone and
              2-butanone, which must be less than two times their respective
              CRQLs.  The concentration of each target compound in the
              instrument blank must be less than its CRQL listed in Exhibit C
              (Volatiles).  The concentration of non-target compounds-in all
              blanks must be less than 2.0 ug/L.

12 .1.6     Corrective Action for Blank Analyses

12.1.6.1      It is the Contractor's responsibility to ensure that method
              interferences caused by contaminants in solvents, reagents,
              glassware, laboratory air, and other sample storage and processing
              hardware that lead to discrete artifacts and/or elevated baselines
              in gas chromatograms, be eliminated.  If a Contractor's blanks
              exceed the criteria in Section 12.1.5.7, the Contractor must
              consider the analytical  system to be out of control.  The source
              of the contamination must be investigated and appropriate
              corrective measures MUST be taken and documented before further
              sample analysis proceeds.

12.1.6.2      Any method blank  or instrument blank that fails to meet the
              technical acceptance criteria must be re-analyzed at no additional
              cost  to USEPA.  Further, all samples processed within the 12-hour
              time  period with  a method blank  or instrument blank that does not
              meet  the blank technical acceptance criteria will require re-
              analysis at no additional cost to USEPA.

12.1.6.3      If the  storage blank does not meet the  technical acceptance
              criteria for blank analyses  in Sections 12.1.5.1 to 12.1.5.6,
              correct  system problems  and  re-analyze  the storage blank.  If the
              storage  blank  does not meet  the  criteria in Section 12.1.5.7, re-
              analyze  the blank to determine whether  the contamination occurred
              during  storage or during analyses.  If  upon re-analysis, the
              storage  blank  meets the  criteria in Section 12.1.5.7, the problem
              occurred during  the analysis and the re-analyzed storage blank
              results  must be  reported.   If upon re-analysis  the storage blank
              did not  meet  the  criteria  in Section 12.1.5.7,  the problem
              occurred during  storage.  The laboratory manager or.his/her
              designee must  address  the problem in the SDG Narrative  and discuss
              the corrective actions implemented to prevent  future occurrences.

 OLC03.2                                 D-46/VOA

-------
                                                        Exhibit D Volatiles -- Section 12
                                                                  Quality Control  (Con't)

              NOTE:  A copy of the storage blank data must be retained by the
              Contractor and be made available for inspection during on-site
              laboratory evaluations.

12.2   Matrix Spike/Matrix Spike Duplicate (MS/MSD)

12.2.1    Summary of MS/MSD

          In  order  to evaluate  the  effects of  the  sample  matrix on  the  method
          used  for  volatile  analysis, USEPA has prescribed  a  mixture  of
          volatile  target  compounds to be spiked  into two aliquots  of a sample,
          and analyzed  in  accordance with the  appropriate method, upon  request.

12.2.2    Frequency of  MS/MSD

12.2.2.1      A MS/MSD shall only be analyzed if requested by the Region
              (through the Sample Management  Office (SMO)) or specified on the
              Traffic Report (TR).   If  requested,  a matrix spike and a matrix
              spike duplicate must be performed for each group of 20 field
              samples in an SDG, or each SDG,  whichever is most frequent.

12.2.2.2      As part of USEPA's Quality Assurance (QA)/Quality Control  (QC)
              program,  water rinsate samples  and/or field/trip blanks (field QC)
              may be delivered to a laboratory for analysis.  The Contractor
              shall not perform MS/MSD  analysis on any of the field QC samples.

12.i'2.2.3      If the USEPA Region requesting  MS/MSD designates a sample to be
  '-"'          used as an MS/MSD, then that sample must be used.  If there is
  -"-          insufficient sample,  less than  the required amount,  remaining to
   *          perform an MS/MSD, then the Contractor shall choose another sample
   ~-          to perform an MS/MSD analysis.   At the time the selection is made,
   '          the Contractor shall notify the Region (through SMO)  that
              insufficient sample was received and identify the USEPA sample
              selected for the MS/MSD analysis.  The rationale for the choice of
              a sample other than the one designated by the Region shall be
              documented in the SDG Narrative.

12.2.2.4      If there is insufficient  sample remaining in any of the samples in
              an SDG to perform an MS/MSD,  then the Contractor shall immediately
              contact SMO to inform them of the problem.  SMO will contact the
              Region for instructions.   The Region will either approve that no
              MS/MSD is.required, or require  that a reduced sample aliquot be
              used for the MS/MSD analysis.   SMO will notify the Contractor of
              the resolution.  The Contractor shall document the decision in the
              SDG Narrative.

12.2.2.5      If it appears that the Region has requested MS/MSD analysis at a
              greater frequency than specified in Section 12.2.2.1,  the
              Contractor shall contact  SMO.   SMO will contact the Region to
              determine which samples should have an MS/MSD performed on them.
              SMO will notify the Contractor  of the Region's decision.  The
              Contractor shall document the decision in the SDG Narrative.  If
              this procedure is not followed,  the Contractor will not be paid
              for MS/MSD analysis performed at a greater frequency than required
              by the contract.

                                         D-47/VOA                                  OLC03.2

-------
Exhibit D Volatiles -- Section 12
Quality Control (Con't)

12.2.2.6      When a Contractor receives only Performance  Evaluation (PE)
              sample(s),  no MS/MSD shall be performed within that  SDG.

12.2.2.7      When a Contractor receives a PE sample  as part of a  larger  SDG,  a
              sample other than the PE sample must  be chosen for the requested
              MS/MSD analysis when the Region did not designate samples to be
              used for this purpose.

12.2.3    Procedure  for  Preparing MS/MSD

12.2.3.1      To prepare MS/MSD samples, add 10 uL  of the  matrix spike solution
              (Section 7.2.3.5) to each of the 25 mL  aliquots of the sample
              chosen for spiking.  Process samples  according to Section 10.2.
              Disregarding any dilutions, this is equivalent to a  concentration
              of 5 ug/L of each matrix spike compound.

12.2.3.2      MS/MSD samples must be analyzed at the  same  concentration as  the
              most concentrated aliquot for which the original sample results
              will be reported.  Sample dilutions must be  performed in
              accordance with Section 10.2.9.  Do not further dilute MS/MSD
              samples to get either spiked or non-spiked analytes  within
              calibration-range.

12.2.4    Calculations for MS/MSD

12.2.4.1      Calculate the concentrations of the matrix spike compounds  using
              the same equations as used for target -compounds  (Equation  6) .
              Calculate the recovery of each matrix spike  compound as follows:
              EQ. 9
                       Hurrlx Spike Recovery • SSR " SR x 100
              Where:

               SSR   =  Spiked sample result.

               SR    =  Sample result.

               SA    =  Spike added.

 12.2.4.2      Calculate the  Relative Percent Difference  (RPD) of the recoveries
              of each compound  in the MS/MSD as follows:

              EQ. 10
                                                         100
 OLC03.2                                 D-48/VOA

-------
                                                        Exhibit D Volatiies -- Section 12
                                                                 .Quality Control  (Con't)

              Where:
               MSR   =  Matrix spike recovery.

               MSDR  =  Matrix spike duplicate recovery.


12.2.5    Technical Acceptance Criteria for MS/MSD

12.2.5.1      If requested,  all MS/MSD must be prepared and .analyzed at the
              frequency described in Section 12.2.2.   All  MS/MSDs must be
              analyzed on a GC/MS system meeting the BFB,  initial and continuing
              calibration technical acceptance criteria,  and the blank technical
              acceptance criteria.

12.2.5.2      The MS/MSD must be  analyzed within the contract holding time.

12.2.5.3      The retention time  shift for each of the internal standards in the
              MS/MSD must be within ±0.33  minutes (20 seconds)  of its retention
            .  time in the most recent continuing calibration standard analysis.

12.2.5.4      The limits for matrix spike compound recovery and RPD are given in
              Table D-6.  As these limits are only advisory, no further action
              by the laboratory is required.   However, frequent failures to meet
              the limits for recovery or RPD warrant investigation by the
              laboratory,  and may result in questions from USEPA.

12.2.5V5      The relative retention time for the DMCs must be within ±0.06 RRT
      *""       units of its standard retention time in the Continuing Calibration
              Standard.

12.2.6    Corrective  Action for  MS/MSD

          Any MS/MSD  that does not meet the  technical acceptance criteria in
          Sections 12.2.5.1 through 12.2.5.3  must be re-analyzed at  no
          additional  cost to USEPA.

12.3   Method Detection Limit (MDL) Determination

12.3.1    Before any  field  samples are analyzed  under this  contract,  the MDL
          for each volatile target compound  shall be determined on each
          instrument  used  for analysis.   The  MDLs must be verified annually
          thereafter  (see Section  12.3.2  for  MDL verification  procedures),
          until the contract expires or is terminated, or after major
          instrument  maintenance.   Major  instrument  maintenance includes, but
          is  not limited to cleaning or replacement  of the  mass spectrometer
          source,  mass filters  (e.g.,  quadrupole, ion trap,  etc.), or electron
          multiplier  (or similar device),  replacement of gas chromatographic
          column,  and replacement  or overhaul of the purge  and trap  device.

12.3.2    To  determine the  MDLs, the Contractor  shall run an MDL study
          following the procedures specified  in  40  CFR Part 136.  The
          Contractor  shall  analyze the MDL samples  on each  instrument used for
          field sample analyses.   MDL  verification  is achieved by analyzing a
          single reagent water blank spiked with each volatile target compound


                                         D-49/VOA                                 OLC03.2

-------
Exhibit D Volatiles -- Section 12
Quality Control  (Con't)

          at a concentration  equal to two times the analytical determined MDL.
          The resulting mass  spectra of each target compound must meet the
          qualitative  identification criteria outlined  in  Sections  11.1.1
          through  11.1.4.3

12.3.3    The determined  concentration of the MDL must  be  less than the CRQL.

12.3.4    All documentation for  the MDL studies shall be maintained at the
          laboratory and  provided to USEPA upon written request.
 OLC03.2                                 D-50/VOA

-------
                                                  Exhibit D Volatiies -- Sections 13 - 16
                                                                       Method Performance

13.0   METHOD PERFORMANCE

       Not applicable.

14.0   POLLUTION PREVENTION

14.1   Pollution prevention encompasses any technique that reduces or
       eliminates the quantity or toxicity of waste at the point of generation.
       Numerous opportunities for pollution prevention exist in laboratory
       operation.  USEPA has established a preferred hierarchy of environmental
       management techniques that places pollution prevention as the management
       option of first choice.  Whenever feasible, laboratory personnel should
       use pollution prevention techniques to address their waste generation.
       When wastes cannot be feasibility reduced at the source, USEPA
       recommends recycling as the next best option.

14.2   For information about pollution prevention that may be applicable to
       laboratories and research institutions, consult Less is Better:
       Laboratory Chemical Management for Waste Reduction, available from the
       American Chemical Society's Department of Government Relations and
       Science Policy, 1155 16th Street, N.W., Washington D.C., 20036, (202)
       872-4477.

15.0   WASTE MANAGEMENT

15.1   USEPA requires that laboratory waste management practices be conducted
       consistent with all applicable rules and regulations.  USEPA urges
       laboratories to protect the air, water, and land by minimizing and
       controlling all releases from hpods and bench operations, complying with
       the letter and spirit of any sewer discharge permits and regulations,
       and by complying with all solid and hazardous waste regulations,
       particularly the hazardous waste identification rules and land disposal
       restrictions.  For further information on waste management, consult The
       Waste Management Manual for Laboratory Personnel, available from the
       American Chemical Society at the address listed in Section 14.2.

16.0   REFERENCES

       Not applicable.
                                         D-51/VOA                                 OLC03.2

-------
Exhibit D Volatiles -- Section 17
Tables/Diagrams/Flowcharts

17.0   TABLES/DIAGRAMS/FLOWCHARTS

                      TABLE D-l  BFB KEY IONS AND ABUNDANCE CRITERIA


              Mass          Ion Abundance Criteria

              50            8.0-40.0  percent of mass  95
              75            30.0-66.0 percent of mass 95
              95            base  peak,  100 percent relative abundance
              96            5.0-9.0 percent of mass 95 (see note)
              173           less  than 2.0 percent of mass  174
              174           50.0  - 120.0 percent of mass 95
              175           4.0 - 9.0 percent of mass 174
              176           93.0  - 101.0 percent of mass 174
              177 .          5.0-9.0 percent of mass 176

           NOTE: All ion abundances must be normalized to m/z 95,  the nominal
           base peak,  even though the ion abundance of m/z 174 may be up to 120
           percent that of m/z 95.
 OLC03.2                                  D-52/VOA

-------
                                         Exhibit D Volatiles - Section 17
                                       Tables/Diagrams/Flowcharts (Con't)
                        TABLE D-2
TECHNICAL ACCEPTANCE CRITERIA FOR INITIAL AND CONTINUING
       CALIBRATION FOR VOLATILE ORGANIC COMPOUNDS
Volatile
Compound
Dichlorodifluoromethane
Chloromethane
Vinyl chloride
Bromomethane
Chloroethane
Trichlorofluoromethane
1 , 1-Dichloroethene
1, l,2-Trichloro-l,2, 2-trif luoroethane
Acetone
Carbon disulfide
Methyl Acetate
Methylene chloride
trans-1, 2-Dichloroethene
Methyl tert -Butyl Ether
1, 1-Dichloroethane
cis-1, 2-Dichloroethene
2-Butanone
Bromochloromethane
Chloroform
1,1, 1-Trichloroethane
Cyclohexane
Carbon tetrachloride
Benzene
1 , 2 -Dichloroethane
Trichloroethene
Methyl cyclohexane
1 , 2 -Dichloropropane
Bromodichloromethane
cis-1, 3-Dichloropropene
4 -Methyl - 2 -pent anone
Toluene
trans-l , 3 -Dichloropropene
1,1,2-Trichloroethane
Tetrachlofoethene
2-Hexanone
Dibromochloromethane
1 , 2 -Dibromoethane
Chlorobenzene
Ethylbenzene
Xylene (total)
Styrene
Bromoform
I sopropy Ibenzene
Minimum
RRF
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
010
010
100
100
010
010
100
010
010
010
010
010
010
010
200
010
010
050
200
100
010
100
400
100
300
010
010
200
200
010
400
100
100
100
010
100
100
500
100
300
300
050
010
Maximum
%RSD
none
none
30.0
30.0
none
none
30.0
none
none
none
none
none
none
none
30.0
none
none
30.0
30.0
30.0
none
30.0
30.0
30.0
30.0
none
none
30.0
30.0
none
30.0
30.0
30.0
30.0
none
30.0
30.0
30.0
30.0
30.0
30.0
30.0
none
Maximum
%Diff
none
none
±30.0
±30.0
none
none
±30.0
none
none
none
none
none
none
none
±30.0
none
none
±30.0
±30.0
±30.0
none
±30. .0
±30.0
±30.0
±30.0
none
none
±30.0
±30.0
none
±30.0
±30.0
±30.0
±30.0
none
±30.0
±30.0
±30.0
±30.0
±30.0
±30.0
±30.0
none
                        D-53/VOA
OLC03.2

-------
Exhibit D Volatiles - Section 17
Tables/Diagrams/Flowcharts  (Con't)
                                         TABLE  D-2
                 TECHNICAL ACCEPTANCE CRITERIA FOR INITIAL AND CONTINUING
                    CALIBRATION FOR VOLATILE ORGANIC COMPOUNDS (Con't)
Volatile
Compound
1,1,2,2 -Tetrachloroethane
1 , 3 -Dichlorobenzene
1 , 4 -Dichlorobenzene
1 , 2 -Dichlorobenzene
1, 2-Dibromo-3-chloropropane
1 , 2 , 4-Trichlorobenzene
1,2,3 -Trichlorobenzene
DEUTERATED
Vinyl Chloride-d3
Chloroethane-d5
1, l-Dichloroethene-d2
2-Butanone-d5
Chloroform-d
1 , 2 -Dichloroethane-d4
Benzene-d6
1 , 2 -Dichloropropane-d6
Toluene-d8
trans-1, 3-Dichloropropene-d4
2-Hexanone-d5
Bromoform-d
1, l,2,2-Tetrachloroethane-d2
l,2-Dichlorobenzene-d4
Minimum
RRF.
0.100
0.400
0.400.
0.400
0.010
0.200
0.200
MONITORING COMPOUNDS
0.010
0.010
0.010
0.010
0.010
0.010
0.010
0.010
0.010
0.010
0.010
0.010
0.010
0.010
Maximum
%RSD
30.0
30.0
30.0
30.0
none
30.0
30.0

none
none
none
none
none
none
none
none
none
none
none
none
none
none
Maximum
%Diff
±30.0
±30.0
±30.0
±30.0
none
±30.0
±30.0

none
none
none
none
none
none
none
none
none
none
none
. none
none
none
 OLC03.2
D-54/VOA

-------
                                                              Exhibit D Volatiles -- Section 17
                                                             Tables/Diagrams/Flowcharts (Con't)

                                              TABLE D-3
                 VOLATILE TARGET  COMPOUNDS  AND  DEUTERATED MONITORING COMPOUND (DMC)
                       WITH  CORRESPONDING INTERNAL  STANDARDS•FOR QUANTITATION
1,4-Difluorobenzene
Chlorobenzene-d5
1,4-Dichlorobenzene-d4
Dichlorodi fluoromethane
Chloromethane
Vinyl Chloride
Bromomethane
Chloroethane
Trichlorofluoromethane
1,1-Dichloroethene
1,1,2-Trichloro-1,2,2-
  trifluoroethane
Acetone
Carbon Disulfide
Methyl Acetate
Bromochloromethane
Methylene Chloride
trans-1,2-Dichloroethene
Methyl tert-Butyl Ether
1,1-Dichl oroe'thane
cis-1,2-Dichloroethene
2-Butanone
Chloroform
1,2-Dichloroethane
Vinyl Chloride-d3  (DMC)
Chloroethane-d5  (DMC)
l,l-Dichloroethene-d2  (DMC)
2-Butanone-d5  (DMC)
Chlorofortn-d  (DMC)
1,1,1-Trichloroethane
Cyclohexane
Carbon Tetrachloride
Benzene
Trichloroethene
Methy1eye1ohexane
1,2-Dichloropropane
Bromodi chloromethane

cis-1,3-Dichloropropene
4-Methyl-2-pent anone
Toluene
trans-1,3-Dichloropropene
1,1,2-Trichloroethane
Tetrachloroethene
2-Hexanone
Dibromochlbromethane
1,2-Dibromoethane
Chlorobenzene
Ethylbenzene
Xylenes  (total)
Styrene
Isopropylbenzene
1,1,2,2-Tetrachloroethane
Benzene-d6  (DMC)
1,2-Dichloropropane-
  d6  (DMC)
Bromoform
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1,2-Dichlorobenzene
1,2-Dibromo-3-chloropropane
1,2,4-Trichlorobenzene
1,2,3-Trichlorobenzene
l,2-Dichlorobenzene-d4 (DMC)

Bromoform-d  (DMC)
l,2-Dichloroethane-d4  (DMC)
trans-1,3-Dichloropropene-
  d4  (DMC)
Toluene-d8  (DMC)
2 -Hexanone-d5  (DMC)
1,1,2,2-Tetrachloroethane-
  d2  (DMC)
                                               D-55/VOA
                                                      OLC03.2

-------
Exhibit D Volatiles -- Section 17
Tables/Diagrams/Flowcharts  (Con't)
                                        TABLE D-4
                    CHARACTERISTIC IONS FOR VOLATILE TARGET COMPOUNDS
Target Compound
Dichlorodifluoromethane
Chloromethane
Vinyl chloride
Bromomethane
Chloroethane
Trichlorofluoromethane
1 , 1 -Dichloroethene
1, 1, 2-Trichloro-l, 2, 2-trif luoroethane
Acetone
Carbon disulfide
Methyl Acetate
Methylene chloride
trans-l, 2 -Dichloroethene
Methyl tert -Butyl Ether
1 , 1-Dichloroethane
cis - 1 , 2 -Dichloroethene
2 -Butanone
Chloroform
Bromochl o rome thane
1,1, 1-Trichloroethane
Cyclohexane
Carbon Tetrachloride
Benzene
1 , 2-Dichloroethane
Trichloroethene
Methyl cyclohexane
1 , 2 -Dichloropropane
Bromodichloromethane
cis-1, 3 -Dichloropropene
4 -Methyl -.2 -pentanone
Toluene
trans-l , 3 -Dichloropropene
1,1, 2-Trichloroethane

Tetrachloroethene
2 -Hexanone
Dibromochloromethane
1 , 2 -Dibromoethane '
Chlorobenzene
Ethylbenzene
Xylene (total)
Styrene
Bromoform
*m/z 43 is used for quantitation of 2-Butanone,
Primary
Quantitation Secondary
. Ion Ion(s)
85
50
62
94
64
101
96
101
43
76
43
84
96 .
73
63
96
43*
83
128
97
56
117
78
62
95
83
63
83
75
43
91
75
97

164
43
129
107
112
91
106
104
173
but m/z 72 must
87
52
64
96
66
103
61, 63
85, 151
58
78
74
49, 86
61, 98
43, 57
65, 83
61, 98
72
85
49, 130,
99, 61
69, 84
119
-
98
97, 132,
55, 98 -
112
85, 127
77
58, 100
92
77
83, 85,
132, 134
129, 131
58, 57,
127
109, 188
77, 114
106
91
78
175, 254
be present for .


















51





130







99,

, 166
100








 positive identification.
 OLC03.2
D-56/VOA

-------
                                        Exhibit D Volatiles  -- Section 17
                                       Tables/Diagrams/Flowcharts  (Con't)
                        TABLE D-4
CHARACTERISTIC IONS FOR VOLATILE TARGET COMPOUNDS  (Con't)


Analyte
Isopropylbenzene
1,1,2 , 2-Tetrachloroethane
1 , 3 -Dichlorobenzene
1 , 4 -Dichlorobenzene
1 , 2 -Dichlorobenzene
1 , 2 -Dibromo-3 -Chloropropane
1,2,4 -Trichlorobenzene
1,2,3 -Trichlorobenzene
Deuterated Monitoring Compounds
Vinyl Chloride-d3
Chloroethane-d5
1, l-Dichloroethene-d2
2 - But anone - d5
Chloroform-d
1, 2-Dichloroethane-d4
Benzene-d6
1 , 2 -Dichloropropane-d6
Toluene-d8
t rans -1,3-Dichl orqpir opene - d4
2 - Hexanone - d5
Bromoform-d
1,1,2,2 -Tetrachloroethane-d2
1, 2-Dichlorobenzene-d4
Internal Standards
1 , 4 -Dichlorobenzene-d4
1 , 4 -Dif luorobenzene
Chlorobenzene-d5
Primary
Quantitation Secondary
Ion Ion(s)
105 120, 77
83 85, 131
146 111, 148
146 111, 148
146 111, 148
75 157, 155
180 182, 145
180 182, 145

65 67
69 71, 51
63 98, 65
46 77
84 86, 47, 49
65 67, 51
84 82, 54, 52
67 65, 46, 42
98 100, 42
79 . 81, 42
63 46
.174 . 172
84 86
152 150

152 115, 150
114 63, 88
117 82, 119
                         D-57/VOA
OLC03.2

-------
Exhibit D Volatiles -- Section 17
Tables/Diagrams/Flowcharts  (Con't)
                                        TABLE D-5  .
                      DEUTERATED MONITORING COMPOUND  RECOVERY LIMITS
                   Compound
                 Percent
             Recovery Limits
                   Vinyl  Chloride-d3
                   Chioroethane-d5
                   1,l-Dichloroethene-d2
                   2-Butanone-d5
                   Chi oro f orm- d
                   1,2-Dichloroethane-d4
                   Benzene-d6
                   1,2-Dichloropropane-d6
                   Toluene-d8
                   trans-1,3-Dichloropropene-d4
                   2-Hexanone-d5
                   Bromoform-d
                   1,1,2,2-Tetrachloroethane-d2
                   1,2-Dichlorobenzene-d4
                  49-138
                  60-126.
                  65-130
                  42-171
                  80-123
                  78-129
                  78-121
                  84-123
                  77-120
                  80-128
                  37-169
                  76-135
                  75-131
                  50-150
           NOTE: The recovery limits for any of the compounds listed above may
           be expanded at any time during the period of performance if USEPA
           determines that the limits are too restrictive.
                                         TABLE D-6
                                 MATRIX SPIKE RECOVERY AND
                            RELATIVE PERCENT DIFFERENCE LIMITS
                   Compound
    % Recovery
                      RPD
1 , 1 -Dichloroethene
Benzene
Trichloroethene
Toluene
Chlorobenzene
61-145
76-127
71-120
76-125
75-130
14
11
14
13
13
 OLC03.2
D-58/VOA

-------
                       EXHIBIT D
METHOD FOR THE ANALYSIS OF LOW CONCENTRATION WATER FOR
            SEMIVOLATILE ORGANIC COMPOUNDS
                         D-l/SV                                  OLC03.2

-------
                           THIS PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                    D-2/SV

-------
                    Exhibit D -- Analytical Methods for Setnivolatiles

                                    Table of Contents

Section                                                                  •            Page


1.0    SCOPE AND APPLICATION	•	5

2.0    SUMMARY OF METHOD	6

3.0    DEFINITIONS  	 6

4.0    INTERFERENCES   	 7

5.0    SAFETY	7

6.0    EQUIPMENT AND SUPPLIES 	 8
       6.1    Glassware	8
       6.2    Gases	8
       6.3    Gas-line Tubing	  . 8
       6.4    Silicon Carbide Boiling Chips 	 8
       6.5    Water Bath	8
       6.6    Balance	8
       6.7    Nitrogen Evaporation Device  	 9
       6.8    pH Meter   	9
       6.9    pH Paper   	9
       6.10   Gas Chromatograph/Mass Spectrometer (GC/MS)	 9

7.0    REAGENTS AND STANDARDS	  12
       7.1    Reagents   	  12
       7.2    Standards	  12

8.0    SAMPLE COLLECTION, PRESERVATION, AND STORAGE 	  16
       8.1    Sample Collection and Preservation  	  16
       .8.2    Procedure for Sample Storage  	  16
       8.3    Procedure for Sample Extract Storage  	  16
       8.4    Records for Sample and Sample Extract Storage  	  16
       8.5    Contract Required Holding Times   	  16

9.0    CALIBRATION AND STANDARDIZATION	17
       9.1    Instrument Operating Conditions   	  17
       9.2    Instrument Performance Check (DFTPP)  	  17
       9.3    Initial Calibration  	  18
       9.4    Continuing Calibration   	  21

10.0   PROCEDURE	24
       10.1   Sample Preparation   	  '24
       10.2   Instrument Analysis of Sample 	  27

11.0   DATA ANALYSIS AND CALCULATIONS	29
       11.1   Qualitative Identification of Target Compounds   	  29
       11.2   Qualitative Identification of Non-Target Compounds   	  30
       11.3   Calculations for Target Compounds  	  31
       11.4   Calculations for Non-Target Compounds .  . .'	34
       11.5   Calculations for Deuterated Monitoring Compounds (DMCs)   	  34
       11.6   Technical Acceptance Criteria for Sample Analysis  	  :  .  .  34
       11.7   Corrective Action for Sample Analysis 	  .  	  35

12.0   QUALITY CONTROL	  36
       12.1   Method Blank	36
       12.2   Matrix Spike/Matrix Spike Duplicate (MS/MSD)   	  37
       12.3   Method Detection Limit (MDL) Determination  '	40
                                          D-3/SV                                  OLC03.2

-------
                    Exhibit D -- Analytical Methods for Semivolatiles





                              Table of Contents (Continued)





                                                                                     Page








13.0   METHOD PERFORMANCE 	  41





14.0   POLLUTION PREVENTION	.41





15.0   WASTE MANAGEMENT  	41





16.0   REFERENCES	  41





17.0   TABLES/DIAGRAMS/FLOWCHARTS	42
 OLC03.2        '                           D-4/SV

-------
                                                     Exhibit D Semivolatiles -- Section 1
                                                                    Scope and Application
1.0    SCOPE AND APPLICATION
1.1    The analytical method that follows is designed to analyze water samples
       containing low concentrations of the semivolatile compounds listed on
       the Target Compound List (TCL)  in Exhibit C.  The majority of the
       samples are expected to be from drinking water and well/ground
       water/aqueous type sources around Superfund sites.   The method is based
       upon the semivolatile method contained in the Contract Laboratory
       Program (CLP) Statement of Work, "Organic Analysis, Multi-Media,  Multi-
       Concentration Analyses".  The analytical method includes the use of
       Deuterated Monitoring Compounds (DMC) for precision and accuracy
       assessment.

1.2    Benzaldehyde, Acetophenone,  Caprolactam, 1,1'-Biphenyl, Atrazine, and
       1,2,4,5-Tetrachlorobenzene have been added to the TCL.
1.3    Problems have been associated with the following compounds analyzed by
       this method:

       •    3,3'-Dichlorobenzidine  and  4-chloroaniline  may be  subject  to
            oxidative losses  during solvent  concentration.

       •    Hexachlorocyclopentadiene is  subject  to thermal decomposition  in  the
            inlet  of the Gas  Chromatograph (GC),  chemical  reactions  in acetone
            solution,  and photochemical decomposition.

       •    N-nitrosodiphenylamine  decomposes in  the gas chromatographic inlet
            forming diphenylamine and,  consequently,  may be detected as
            diphenylamine.

       •    Due  to the lower  quantitation limits  required  by this method,  extra
            caution must be exercised when identifying  compounds.
                                          D-5/SV                                  OLC03.2

-------
'Exhibit D Semivolatiles --  Sections 2 & 3
 Summary of Method

 2.0    SUMMARY OF METHOD

 2.1    A one liter aliquot  of sample is acidified to pH 2.0  and extracted with
        methylene chloride using a- continuous liquid-liquid extractor.
        Separator/ funnel extraction is not permitted.   The methylene chloride
        extract is dried with sodium sulfate and concentrated to a volume of 1.0
        milliliter (mL).  The extract is injected onto a Gas  Chromatograph (GC)
        capillary column. The GC is temperature programmed to separate the
        semivolatile compounds,  which are then detected with  a Mass Spectrometer
        (MS) .

 2.2    Deuterated Monitoring Compounds (DMCs)  and internal standards are added
        to all samples,  standards,  requested Matrix Spike/Matrix Spike
        Duplicate(s)  (MS/MSD) and blanks.  The target compounds and DMCs are
        identified in the samples and blanks by analyzing standards that contain
        all target compounds, DMCs, and internal standards under the same
        conditions and comparing resultant mass spectra and GC retention times.
        A Relative Response  Factor  (RRF) is established for each target compound
        and DMC during the initial and continuing calibrations by comparing the
        mass spectra response from the Extracted Ion Current  Profile (EICP) for
        the primary quantitation ion produced by that compound to the mass
        spectra response for the primary quantitation ion produced by the
        associated internal  standard compound.  Each identified target compound
        and DMC is quantitated by comparing the instrument response for the
        compound in the sample, standard, requested MS/MSD or blank with the
        instrument response  of the associated internal standard, while taking
        into account the RRF from the most recent mid-point calibration, the
        sample volume, and any sample dilutions.

 2.3    Non-target compounds are identified by comparing the  resultant mass
        spectra from the non-target compounds to mass spectra contained in the
        NIST/EPA/NIH  (May 1992 release or later) and/or Wiley  (1991 release or
        later) or equivalent mass spectral library.  Non-target compounds are
        quantitated by comparing the mass spectra response from the
        Reconstructed Ion Chromatogram  (RIC) for the non-target compound peaks
        to the mass spectra response produced by the nearest  internal standard.
        An RRF of 1 is assumed.

 3.0    DEFINITIONS

        See  Exhibit G for a complete list of definitions.
  OLC03.2                                    D-6/SV

-------
                                                Exhibit D Semivolatiles -- Sections 4 & 5
                                                                            Interferences
4 .0    INTERFERENCES
       Contaminants in solvents,  reagents,  glassware,  and other sample
       processing hardware may cause method interferences such as discrete
       artifacts and/or elevated baselines in the Reconstructed Ion
       Chromatogram (RIC)  profiles.  All of these materials must be routinely
       demonstrated to be free from interferences under the conditions of the
       analysis by running laboratory method blanks.   Matrix interferences may
       be caused by contaminants that are coextracted from the sample.  The
       extent of matrix interferences will vary considerably from source to
       source.
5.0    SAFETY
       The toxicity or carcinogenicity of each reagent used in this method has
       not been precisely determined; however, each chemical should be treated
       as a potential health hazard.  Exposure to these reagents should be
       reduced to the lowest possible level.   The laboratory is responsible for
       maintaining a current awareness file of OSHA regulations regarding the
       safe handling of the chemicals specified in this method.  A reference
       file of data handling sheets should be made available to all personnel
       involved in these analyses.  Specifically, concentrated sulfuric acid
       presents some hazards and is moderately toxic and extremely irritating
       to skin and mucous membranes.  Use these reagents in a fume hood
       whenever possible and if eye or skin contact occurs flush with large
     3 volumes of water.  Always wear safety glasses or a shield for eye
       protection, protective clothing, and observe proper mixing when working
       with these reagents.
                                          D-7/SV                                  OLC03.2

-------
Exhibit D Semivolatiles -- Section 6
Equipment and Supplies

6.0    EQUIPMENT AND SUPPLIES

       Brand names, suppliers, catalog and part numbers are for illustrative
       purposes only.  No endorsement is implied.  Equivalent performance may
       be achieved using equipment and supplies other than those specified
       here, but demonstration of equivalent performance meeting the
       requirements of this Statement of Work is the responsibility of the
       Contractor.  The Contractor shall document any use of alternate
       equipment or supplies in the SDG Narrative.

6.1    Glassware

6.1.1     Continuous liquid-liquid extractors  -  Equipped with PTFE  or glass
          connecting joints  and stopcocks  requiring no lubrication
           (Hershberg-Wolf  Extractor-Ace  Glass  Company,  Vineland, NJ, P/N
          6841-10 or equivalent)  or Hydrophobia  Membrane-based Extractor
           (Accelerated One Step™ Extractor, Corning series 3195 or
          equivalent).

6.1.2     Drying column -  19 millimeter  (mm) ID  chromatographic column  with
          coarse frit (substitution of a small pad of  Pyrex pre-extracted glass
          wool for the frit  will prevent cross contamination of sample
          extracts).

6.1.3     Kuderna-Danish Apparatus

6.1.3.1       Concentrator tube - Kuderna-Danish, 10 milliliter  (mL),  graduated
               (Kontes, Vineland, NJ, K-570050-1025 or equivalent).

6.1.3.2       Evaporation flask - Kuderna-Danish, 500 mL (Kontes K-570001  -0500
              or  equivalent).  Attach to concentrator tube with springs.

6.1.3.3       Snyder column  - Kuderna-Danish, Three-ball macro  (Kontes K-50300-
              0121 or equivalent).

6.1.3.4       Snyder column  - Kuderna-Danish, Two-ball micro  (Kontes K-569001-
              0219 or equivalent).

6.1.4     Vials - Amber glass,  2 mL capacity with PTFE-lined screw-cap.

6.1.5      Syringes - 0.2mL, O.SmL, 10  mL volumes with Luerlock.

6.1.6     Micro-syringes - 10 microliter  (uL)  and larger,  0.006 inch  (0.15 mm)
           ID needle.

6.2    Gases  -  Helium,  Nitrogen,  ultra pure grade.

6.3    Gas-line Tubing -  Stainless  steel or copper tubing.

6.4    Silicon Carbide Boiling  Chips - Approximately 10/40  mesh.  Heat  to 400°C
       for  30 minutes or Soxhlet  extract with  methylene chloride.   PTFE boiling
       chips  solvent rinsed  prior to use are acceptable.

6.5    Water  Bath -  Heated,  with concentric ring cover, capable of  temperature
       control.   To prevent  the release  of solvent fumes into the laboratory,
       the  bath must be used in a hood.

6.6    Balance - Analytical, capable of  accurately weighing ±0.0001 grams (g).
       The  balances must be  calibrated with class S weights or known reference
       weights once per each 12-hour workshift.   The balances must  be
       calibrated with class S  weights at  a minimum of once a month. The
       balances must also be annually checked  by a certified technician.


.OLC03.2                    .                D-8/SV

-------
                                                     Exhibit D Semivolatiles -- Section 6
                                                           Equipment and Supplies  (Con't)

6.7    Nitrogen Evaporation Device - Equipped with a water bath that can be
       maintained at 35°C to 40°C.  To prevent the release of solvent fumes
       into the laboratory, the nitrogen evaporation device must be used in a
       hood.   The N-Evap by Organomation Associates, Inc. South Berlin,  MA (or
       equivalent) is suitable.

6.8    pH Meter - With a combination glass electrode, calibrated according to
       manufacturer's instructions.  The pH meter shall be calibrated before
       each use.

6.9    pH Paper - Including narrow range capable of measuring a pH of 2.

6.10   Gas Chromatograph/Mass Spectrometer (GC/MS)

6.10.1    Gas  Chromatograph  -  The gas  chromatograph system must be  capable of
          temperature programming and  have a  flow controller that maintains a
          constant  column  flow.rate throughout the  temperature program.  The
          system  must be suitable for  splitless  injection  and have  all required
          accessories including syringes, analytical columns, and gases.  All
          GC carrier gas lines must be constructed  from stainless steel or
          copper  tubing.   Non-polytetrafluoroethylene  (PTFE) thread sealants,
          or flow controllers  with rubber components are not to be  used.

6.10.2    Gas  Chromatography Column -  Minimum length 30 meters  (m)  x 0.25
          millimeter  (mm)  ID (or  0.32  mm) bonded-phase  silicon coated fused
          silica  capillary column DB-5 (J&W Scientific); RTx-5  (Restek); SPB-5
           (Supelco); AT-5  (Alltech); HP-5  (Hewlett-Packard); CP-Sil 8CB
           (Chrompack);  007-2 (Quadrex);  BP-5  (SGE);  or  equivalent.   Note that
          this is a minimum  requirement for column  length.  Longer  columns may
      •••*   be used.  Although a film thickness of 1.0 micron is recommended
      ~    because of its larger capacity, a film thickness of 0.25  micron may
          be used.  A description of the GC column  used for analysis shall be
      3?   provided  in the  SDG  Narrative.

6.10.2.1      A capillary column is considered equivalent if:

              •   The column does not  introduce  contaminants which  interfere
                  with  the identification and quantitation of the compounds
                  listed in  Exhibit C  (Semivolatiles).

              •   The analytical  results generated  using the column meet the
                  initial  and  continuing calibration technical acceptance
                  criteria listed in the SOW, and the Contract Required
                  Quantitation Limits  (CRQLs) listed in Exhibit C
                   (Semivolatiles).

              •   The column can  accept up to 120 nanograms  (ng) of each
                  compound listed in Exhibit  C  (Semivolatiles) without becoming
                  overloaded.

              •   The column provides  equal or better resolution of the
                  compounds  listed in  Exhibit C  (Semivolatiles) than the columns
                  listed in  Section 6.10.2.

6.10.2.2      As applicable,  follow manufacturer's instructions for use of its
              product.

6.10.2.3      The Contractor must maintain .documentation that the alternate
              column met the criteria in Section 6.10.2.1.   The minimum
              documentation is as follows:
                                          D-9/SV                                  OLC03.2

-------
Exhibit D Semivolatiles -- Section 6
Equipment and Supplies (Con't)

6.10.2.3.1       Manufacturer provided information concerning the performance
                 characteristics of the column.

6.10.2.3.2       Reconstructed Ion Chromatograms  (RICs) and data system reports
                 generated on the GC/MS used for CLP analyses:

                 •    From blanks which demonstrate that there are no
                      contaminants which interfere with the semivolatile analysis
                      when using the alternate column;

                 •    For initial calibration standards analyzed using the
                      alternate column,-

                 •    For continuing calibration standards analyzed using the
                      alternate column.
6.10.2.4      Based on the Contractor generated data described in Section
              6.10.2.3.2, the Contractor must complete a written comparison and
              review signed by the Laboratory Manager certifying that:

              •    The column performance meets  the  technical  acceptance criteria
                   in Sections 9.3.5  and 9.4.5.

              •    The low point initial calibration standard  analysis  has
                   adequate sensitivity to meet  the  semivolatile CRQLs.

              •    The high point initial calibration standard analysis was not
                   overloaded.

              •    The column does not introduce contaminants  which interfere
                   with the identification and/or quantitation of compounds
                   listed in Exhibit  C (Semivolatiles).

6.10.2.5      The documentation must be made available to USEPA during on-site
              laboratory evaluations or sent to USEPA upon request of the
              Contract Laboratory Program Project Officer (CLP PO) or the
              Organic Program Manager at Analytical Operations/Data Quality
              Center  (AOC).

6.10.2.6      PACKED COLUMNS  CANNOT BE USED.

6.10.3     Mass Spectrometer - The mass spectrometer must be capable of scanning
           from 35 to 500 atomic mass units (amu)  every 1 second or less,
           utilizing 70 volts (nominal)  electron energy in the electron impact
           ionization mode and producing a mass spectrum which meets all the
           decafluorotriphenylphosphine (DFTPP)  GC/MS performance check
           technical acceptance criteria (Table D-l) when 50 ng of DFTPP is
           injected through the GC inlet.  To ensure sufficient precision of
           mass spectral data, the MS scan rate must allow acquisition of at
           least five spectra while a sample compound elutes from the GC.  The
           GC/MS system must be in a room whose atmosphere is demonstrated to be
           free of all potential contaminants which will interfere with the
           analysis.  The instrument must be vented outside the facility or to a
           trapping system which prevents the release of contaminants into the
           instrument room.

 6.10.4     GC/MS interface - any GC/MS interface which provides acceptable
           sensitivity at CRQLs.  GC/MS interfaces constructed of all-glass or
           glass-lined materials are recommended.  Glass can be deactivated by
           silanizing with dichlorodimethylsilane.
 OLC03.2                                  D-10/SV

-------
                                                     Exhibit D Semivoiatiies -- Section 6
                                                           Equipment and Supplies  (Con't)

6.10.5    Data system  - a computer system must be interfaced to the mass
          spectrometer that allows the continuous acquisition and storage on
          machine-readable media, of all mass spectra obtained throughout the
          duration of  the chromatographic program.  The computer must have
          software that allows  searching any GC/MS data file for ions of a
          specified mass and plotting such ion abundance versus time or scan
          number.  This type of plot is defined as an Extracted Ion Current
          Profile  (EICP).  Software must also be available that allows
          integrating  the abundance in any EICP between specified time or scan
          number limits.  Also, for the non-target compounds, software must be
          available that allows for the comparison of sample spectra against
          reference library spectra.  The NIST/EPA/NIH  (May 1992 release or
          later) and/or Wiley  (1991 release or later), or equivalent mass
          spectral library shall be used as the reference library.  The
          operational  data system must be able to flag all data files that have
          been edited  manually  by laboratory personnel.

6.10.6    Magnetic tape storage device - must be capable of recording data and
          suitable for long-term, off-line storage of GC/MS data.
                                          D-ll/SV                                 OLC03.2

-------
Exhibit D Semivolatiles -- Section 7
Reagents, and Standards

7.0    REAGENTS AND STANDARDS

7.1    Reagents

       Reagents shall be dated with the receipt date and used on a first-in,
       first-out basis.  The purity of the reagents shall be verified before
       use.

7.1.1      Reagent water  -  defined as  water  in which no  semivolatile  target
           compound  is  observed at or  above  the  Contract Required  Quantitation
           Limit  (CRQL) listed in Exhibit  C  for  that compound  and  in  which no
           non-target compound is observed at or above 10 micrograms  per  liter
           (ug/L).

7.1.1.1       Reagent water may be generated by passing tap water through a
              carbon filter bed containing about 453 grams (g) (1 Ib) of
              activated carbon  (Calgon Corp.,  Filtrasorb-300 or equivalent).

7.1.1.2       Reagent water may be generated using a water purification system
               (Millipore Super-Q or equivalent).

7.1.2      Solvents  - Acetone, methanol, methylene chloride, isooctane,  2-
           propanol,  toluene.   Pesticide quality or equivalent 1

7.1.3      Sodium sulfate - (ACS) Granular or powdered,  anhydrous  (J.T.  Baker
           anhydrous powder,  catalog #73898, J.T.  Baker  anhydrous  granular
           #3375,  or equivalent).  Purify  by heating at  400°C  for  four hours in
           a shallow tray,  cool in a desiccator, and store  in  a  glass bottle.
           CAUTION:  An  open container of sodium  sulfate  may become contaminated
           during storage in laboratory.

7.1.4      Sulfuric  acid solution (1:1)  -  slowly add 50  milliliters (mL)  of
           concentrated H2SO4  (Sp. Gr. 1.84; 36N)  to 50 mL of  reagent water.

7.2    Standards

       The Contractor  must provide  all standard solutions to  be used with this
       contract.  These  standards may be  used only after they have been
       certified according to the procedure in  Exhibit  E.  The Contractor must
       be  able to verify that the standards are certified.  Manufacturer's
       certificates of analysis  must  be retained  by the Contractor and
       presented upon  request.

7.2.1      Stock Standard Solutions

           Stock standard solutions may be purchased or prepared using the
           following procedure.

 7.2.1.1        Accurately weigh about  0.0100 g of  pure material.   Dissolve the
               material in  methylene chloride or another suitable  solvent and
               dilute to  volume in a 10 mL volumetric flask.   Larger volumes may
               be used  at the  convenience  of the analyst.

 7.2.1.2        When  compound purity  is assayed to  be 97  percent or greater,  the
               weight may be used without  correction to  calculate  the
               concentration of the  stock  solution.  If  the compound purity  is
               assayed  to be less than 97  percent,  the weight  must be corrected
               when  calculating the  concentration  of the stock solution.

 7.2.1.3        Fresh stock  standards must  be prepared once every twelve months,
               or  sooner,  if standards have degraded or  concentrated.  Stock
               standards  must  be checked for signs of degradation  or


 OLC03.2                                   D-12/SV

-------
                                                     Exhibit D Semivolatiles -- Section 7
                                                           Reagents and Standards  (Con't)

              concentration just prior to preparing secondary dilution and
              working standards from them.

7.2.2     Secondary Dilution Standards

7.2.2.1       Using stock standards,  prepare secondary dilution standards in
              methylene chloride that contain the compounds of interest either
              singly or mixed together.

7.2.2.2       Fresh secondary dilution standards must be prepared once every
              twelve months,  or sooner,  if standards have degraded.or
              concentrated.  Secondary dilution standards must be checked for
              signs of degradation or concentration just prior to preparing
              working standards from them.

7.2.3     Working  Standards

7.2.3.1       Deuterated Monitoring Compound (DMC)  Standard Spiking Solution

7.2.3.1.1        Prepare a DMC  standard spiking  solution that contains the
                 following compounds at concentrations  shown in methanol:

                                                             Concentration
                  DMC                                              * g/mL
                  Phenol-d5                                           40
                  bis-(2-Chloroethyl)ether-d8                         40
                  2-Chlorophenol-d4                                   40
                  4-Methylphenol-d8                                   40
                  Nitrobenzene-d5                                     40
                  2-Nitrophenol-d4                                    40
                  2,4-Dichlorophenol-d3                               40
     •~     •       4-Chloroaniline-d4                                  40  .
     Vu           Dimethylphthalate-d6                                40
                  Acenaphthylene-d8                                   40
                  4-Nitrophenol-d4                                    40
                  Fluorene-dlO                                        40
     **-.           4,6-Dinitro-methylphenol-d2                         40
                  Anthracene-dlO                                      40
                  Pyrene-dlO                                          40
                  Benzo(a)pyrene-dl2                                  40

7.2.3.1.2         DMC  standards are added to all samples, blanks, requested
                  Matrix Spike/Matrix Spike Duplicate(s) (MS/MSD), and
                  calibration solutions.  The DMC standard spiking solution must
                  be prepared every twelve months or sooner if the solution has
                  degraded or concentrated.

7.2.3.2       Matrix Spiking Solution

7.2.3.2.1        The matrix spiking solution consists of the following:

                 Bases/Neutrals                       Acids
                 Acenaphthene                          Pentachlorophenol
                 2,4-Dinitrotoluene                    Phenol
                 Pyrene                                2-Chlorophenol
                 N-Nitroso-di-n-propylamine            4-Chloro-3-methylphenol
                                                       4-Nitrophenol

7.2.3.2.2        Prepare a spiking solution that contains'each of the
                 base/neutral compounds above at 20 micrograms per milliliter
                  (• g/mL)  in methanol and the acid compounds  at 80 • g/mL in
                 methanol.


                                          D-13/SV                                 OLC03.2

-------
Exhibit D Semivolatiles -- Section 7
Reagents and Standards  (Con't)

7.2.3.2.3        The matrix spiking solution must be prepared every twelve
                 months or sooner if the solution has degraded or concentrated.

7.2.3.3       Instrument Performance Check Solution--DFTPP

              Prepare a 50 nanograms per microliter (ng/uL) solution of
              decafluorotriphenylphosphine (DFTPP)  in methylene chloride.  The
              DFTPP solution must be prepared fresh once every twelve months or
              sooner if the solution has degraded or concentrated.

7.2.3.4       Initial and Continuing Calibration Solutions

7.2.3.4.1        Five  initial calibration  standard  solutions are required for
                 all target compounds  and  DMCs.  Standard concentrations of  5,
                 10, 20,  50,  and 80 ng/uL  are required  for the DMCs and  all  but
                 seven of the target compounds.  The  seven compounds:  2,4-
                 dinitrophenol,  2,4,5-trichlorophenol,  2-nitroaniline, 3-
                 nitroaniline,  4-nitroaniline, 4-nitrophenol, and 4,6-dinitro-2-
                 methylphenol require  calibration at  20,  50, 80, 100,  and 120
                 ng/uL.

7.2.3.4.2        To prepare a calibration  standard  solution, add an appropriate
                 volume of secondary dilution standard  to methylene chloride in
                 a volumetric flask.   Dilute to  volume  with methylene  chloride.

7.2.3.4.3        The 20 ng/uL initial  calibration  solution (80 ng/uL  for the
                 seven compounds listed in Section  7.2.3.4.1)  is the  continuing
                 calibration  solution.

7.2.3.4.4        The five initial calibration solutions must be prepared fresh
                 before use.   The continuing calibration standard  solution  must
                 be prepared  weekly or sooner if the  solution  has  degraded  or
                 concentrated.

7.2.3.5       Internal  Standard  Spiking Solution

7.2.3.5.1        Prepare an internal  standard spiking solution in  methylene
                 chloride or  another suitable solvent that contains  1,4
                 dichlorobenzene-di, naphthalene-d8, acenaphthene-d10,
                 phenanthrene-d10, chrysene-d12,  and perylene-d^ at 2000 ng/uL.
                  It may be necessary to use 5 to 10 percent toluene  in this
                  solution and a few minutes of  ultrasonic mixing  in order  to
                 dissolve all the constituents.

                 NOTE: For automated systems using an injection volume of  less
                  than 10 uL,  the internal standard solution may need to be
                 prepared at  a different concentration.

 7.2.3.5.2         The internal standard spiking  solution must be prepared every
                  twelve months or sooner if the solution has degraded or
                  concentrated.

 7.2.4     Ampulated Standard Extracts

           Standard solutions purchased from a chemical supply house as
           ampulated extracts in glass vials may be retained and used until the
           expiration date provided by the manufacturer.   If no manufacturer's
           expiration date is provided, the standard solutions as ampulated
           extracts may be retained and used for 2 years from the preparation
           date.  Standard solutions prepared by the Contractor which are
           immediately ampulated in glass vials may be retained for 2 years from
           preparation date.  Upon breaking the glass seal, the expiration times


 OLC03.2                                   D-14/SV

-------
                                                     Exhibit D Setnivoiatiies  -- Section 7
                                                           Reagents and Standards  (Con't)

          listed in Sections 7.2.1 to 7.2.3.5 will apply.  The Contractor  is
          responsible for assuring that the integrity of the standards has not
          degraded  (Section 7.2.5.5).

7.2.5     Storage of Standard Solutions

7.2.5.1       Store  the  stock  and secondary standard solutions  at 4°C (±2°C)  in
              PTFE-lined screw-cap amber bottles.

7.2.5.2       Store  the  working standard solutions at 4°C (±2°C)  in PTFE-lined
              screw-cap  amber  bottles.

7.2.5.3       Protect  all standards from light.

7.2.5.4       Samples,  sample  extracts,  and standards must'be stored separately.

7.2.5.5       The Contractor is responsible for maintaining the integrity of
              standard solutions and verifying prior to use.   Storage of
              standard solutions in the freezer may cause some  compounds to
              precipitate.   This means at the minimum,  the standards must be
              brought  to room temperature prior to use,  checked for losses, and
              checked  that all components have remained in solution.  Additional
              steps  may be necessary to ensure all components are in solution.

7.2.6     Temperature Records for Storage  of  Standards

7.2.6.1       The temperature  of all standard storage refrigerators shall be
              recorded daily.

7.2.6.2,"      Temperature excursions shall be noted and appropriate corrective
       is-      actions  shall be taken to correct problems,  when required.
      .&•
7.2.6.3.       Corrective action Standard Operating Procedures (SOPs) shall be
              posted on the refrigerators.
                                          D-15/SV                                 OLC03.2

-------
Exhibit D Semivolatiles -- Section 8
Sample Collection, Preservation and Storage

8.0    SAMPLE COLLECTION, PRESERVATION, AND STORAGE

8.1    Sample Collection and Preservation

8.1.1     Water  samples may  be collected  in  1  liter  (L)  (or  1 quart) amber
          glass  containers and fitted  with screw-caps  lined  with  PTFE.   If
          amber  containers are not available,  the  samples  should  be protected
          from light.  The specific requirements for site  sample  collection  are
          outlined by the Region.

8.1.2     All  samples must be  iced or  refrigerated at  4°C  (±2°C)  from  the time
          of collection  until  extraction.

8.2    Procedure for Sample Storage

8.2.1     The  samples must be  protected from light and refrigerated at 4°C
           (±2°C)  from the  time of receipt until  60 days after delivery of a
          complete,  reconciled data package  to USEPA.   After 60 days the
          samples may be disposed of  in a manner that  complies with all
          applicable regulations.

8.2.2     The  samples must be  stored  in an atmosphere  demonstrated to  be free
          of all potential  contaminants.

8.3    Procedure  for  Sample  Extract Storage

8.3.1     Sample extracts  must be protected  from light and stored at less than
          4°C  (±2°C)  until  365 days  after delivery of  a reconciled, complete
          data package to USEPA.

8.3.2     Samples, sample extracts,  and standards  must be  stored  separately.

8.4    Records for Sample and  Sample Extract Storage

8.4.1     The  temperature of all sample and  sample extract storage
          refrigerators shall  be recorded daily.

8.4.2     Temperature excursions shall be noted and  appropriate  corrective
          actions shall be taken to correct  problems,  when required.

8.4.3      Corrective action SOPs shall be posted on  the refrigerators.

8.5    Contract  Required Holding  Times

8.5.1      Extraction of water samples by continuous  liquid-liquid procedures
           shall  be started within 5 days of  Validated Time of  Sample Receipt
           (VTSR) .

           NOTE:  Separatory funnel extraction procedures are not  permitted.

8.5.2     As part of USEPA1s QA program,  USEPA may provide Performance
           Evaluation  (PE)  samples as standard extracts which the Contractor is
           required to prepare per the instructions provided by USEPA.   PE
           samples must be 'extracted and analyzed concurrently with the samples
           in the SDG.  The extraction holding time  (5 days after VTSR) does not
           apply to PE samples received as standard extracts.

8.5.3      Extracts of water samples must be analyzed within 40 days following
           extraction.
 OLC03.2                                 ' D-16/SV

-------
                                                     Exhibit D Semivolatiles -- Section 9
                                                          Calibration and Standardization

9.0    CALIBRATION AND STANDARDIZATION

9.1    Instrument Operating Conditions

9.1.1     Gas Chromatograph  (GC)

9.1.1.1       The following are the gas chromatographic analytical conditions.
              The conditions are recommended unless otherwise noted.

               Initial Column Temperature Hold       40°C for 4 minutes
               Column  Temperature Program            40-270°C at  10°C/min.
               Final Column Temperature Hold         270°C for 3  minutes after all
                                                     compounds listed in Exhibit C
                                                      (Semivolatiles) have eluted
                                                      (required)
               Injector Temperature                  250-300°C
               Transfer Line Temperature             250-300°C
               Source  Temperature          '         According to manufacturer's
                                                     specifications
               Injector                              Grob-type, splitless
               Sample  Volume                         1 • L
               Carrier Gas                           Helium at 30 cm/sec

9.1.1.2       Optimize GC conditions for analyte separation and sensitivity.
              Once optimized, the same GC conditions must be used for the
              analysis of all standards, samples,  blanks,  matrix spikes,  and
              matrix spike duplicates, if required.

9.1.2     Mass  Spectrometer (MS)

          The  following are the required mass spectrometer analytical
      ''••'  .conditions:

           Electron Energy      70 volts (nominal)
           Mass Range            35 to 500 amu
           Scan Time             Not to exceed 1 second per scan
           lonization  Mode      El

9.2    Instrument Performance Check  (DFTPP)

9.2.1     Summary of  Instrument Performance Check

          The  GC/MS system must be tuned to meet  the manufacturer's
          specifications,  using a suitable calibrant such as perfluoro-tri-n-
          butylamine  (FC-43)  or perfluorokerosene (PFK).  The mass calibration
          and  resolution of the GC/MS system are  verified by the  analysis of
          the  instrument performance check solution (Section 7.2'.3.3).   Prior
          to the analysis of any samples (including requested Matrix
          Spike/Matrix Spike Duplicate (MS/MSD) and PE Samples) blanks and
          calibration standards,  the Contractor must establish that  the  GC/MS
          system meets the mass spectral ion abundance criteria for  the
          instrument  performance check solution containing decafluoro-
          triphenylphosphine (DFTPP).

9.2.2     Frequency of Instrument Performance Check

9.2.2.1       The instrument performance check solution must be analyzed once at
              the beginning of each 12-hour period during which samples,  blanks
              or standards are analyzed.
                                          D-17/SV                                 OLC03.2

-------
Exhibit D Semivolatiles -- Section 9
Calibration and Standardization  (Con't)

9.2.2.2       The 12-hour time period for a instrument performance check and
              standards calibration (initial or continuing calibration criteria)
              begins at the moment of injection of the DFTPP analysis that the
              laboratory submits as documentation of a compliant instrument
              performance check.  The time period ends after 12-hours have
              elapsed according to the system clock.

9.2.3      Procedure for Instrument  Performance Check

           The analysis of the instrument performance check solution may be
           performed as an injection of up to 50 nanograms (ng)  of DFTPP into
           the GC/MS or by adding 50 ng of DFTPP to a calibration standard
           (Section 7.2.3.4.3)  and analyzing the calibration standard.

9.2.4      Technical Acceptance Criteria  for Instrument Performance Check

9.2.4.1       The GC/MS system must be tuned at the frequency described in
              Section 9.2.2.

9.2.4.2       The abundance criteria listed in Table D-l must be met for a 50 ng
              injection of DFTPP.  The mass spectrum of DFTPP must be acquired
              in the following manner: three scans  (the peak apex scan and the
              scans immediately preceding and following the apex) are acquired
              and averaged.  Background subtraction is required, and must be
              accomplished using a single scan acquired no more than.20 scans
              prior to the beginning of the elution of DFTPP.  Do not subtract
              part of the DFTPP peak.

              NOTE: All subsequent standards,  samples, and 'blanks associated
              with  a DFTPP analysis must use the  identical mass spectrometer
              instrument  conditions.

9.2.5      Corrective Action for Instrument Performance Check

9.2.5.1        If the GC/MS performance check technical acceptance criteria are
              not met,  re-tune  the GC/MS  system.   It may  be necessary to clean
              the  ion  source,  clean quadrupoles,  or  take  other  actions  to
              achieve  the technical acceptance  criteria.

9.2.5.2       GC/MS performance  check  technical  acceptance  criteria MUST be met
              before any  standards,  samples, and required blanks  are analyzed.
              Any  standards,  samples,  and required blanks analyzed when GC/MS
              performance check technical acceptance  criteria have not been met
              will  require re-analysis at no additional cost to USEPA.

9.3    Initial  Calibration

9.3.1      Summary of Initial Calibration

           Prior to the analysis of samples and required blanks and after GC/MS
           performance check technical acceptance criteria have been met, each
           GC/MS system must be initially calibrated at a minimum of five
           concentrations  (Section 7.2.3.4) to determine  instrument sensitivity
           and the linearity of GC/MS response- for the semivolatile target and
           Deuterated Monitoring Compounds  (DMCs).

 9.3.2      Frequency of Initial Calibration

 9.3.2.1        Each GC/MS system must be  initially calibrated upon award of the
               contract,  whenever the Contractor takes corrective  action which
               may  change or affect  the initial calibration criteria (e.g., ion
               source cleaning or repair,  column replacement, etc.'),  or  if  the


 OLC03.2                                   D-18/SV

-------
                                                     Exhibit D Semivolatiles -- Section 9
                                                  Calibration and Standardization  (Con't)

              continuing calibration technical acceptance criteria have not been
              met.

9.3.2.2       If time still remains in the 12-hour time period after meeting the
              technical acceptance criteria for the initial calibration,  samples
              and blanks may be analyzed.   It is not necessary to analyze a
              continuing calibration standard within this 12-hour time period,
              if the initial calibration standard that  is the same concentration
              as the continuing calibration standard meets the continuing
              calibration technical acceptance criteria.  Quantitate all sample
              and blank results against the initial calibration standard that is
              the same concentration as the continuing  calibration standard
              (Section 7.2.3.4.3).  Compare quality control criteria such as
              internal standard area response change and retention time shift to
              the initial calibration standard that is  the same concentration as
              the continuing calibration standard.

9.3.3     Procedure  for Initial  Calibration

9.3.3.1       Set-up .the GC/MS system per the requirements of Section 9.1.

9.3.3.2       All standard/spiking solutions must be allowed to warm to ambient
              temperature (approximately 1 hour) before preparation or analysis.

9.3.3.3       Prepare five calibration standards containing all the semivolatile
              target and DMCs at the concentrations described in Section
              7.2.3.4.1.

9.3.3.4       Add 10 microliters  (uL) of the internal standard spiking solution
     •*        (Section 7.2.3.5) to 1.0 milliliters (mL) of each of the five
              calibration standards for a concentration of 2.0 nanograms per
              microliter (ng/uL)  for each internal standard compound.

9.3.3.5       Tune the GC/MS system to meet the technical acceptance criteria in
     -?"       Section 9.2.4 for DFTPP.

9.3.3/6       Analyze each calibration standard by injecting 1.0 uL of standard.

9.3.4     Calculations for Initial  Calibrations

9.3.4.1       Calculate Relative Response Factors (RRF) for each semivolatile
              target compound and DMC using Equation 1.  See Table D-2 to
              associate semivolatile target and deuterated monitoring compounds
              with the proper internal standard.  See Table D-3 for primary
              quantitation ions to be used for each semivolatile target
              compound, DMC, and internal standard.

              NOTE: Unless otherwise stated the area response is that of the
              primary quantitation ion.

              EQ. 1
                                                  c.
                                          D-19/SV                                 OLC03.2

-------
Exhibit D Semivolatiles -- Section 9
Calibration and Standardization (Con't)

              Where:


              AX =     Area of  the characteristic  ion  for the compound to be
                       measured.

              Ais  =     Area of  the characteristic  ion  for the specific internal
                       standard (Table D-2).

              Cls  =     Amount of  the internal  standard injected  (ug).

              Cx  =    Amount of  the compound  to be measured injected  (ug).

9.3.4.2       Calculate the mean Relative Response Factor  for each compound
              using Equation 2.

              EQ.  2
                                       X-
              Where:

              Xi =     Each individual value.

              •  =     The mean of n values.

              n  =      The total number of values.

 9.3.4.3       Calculate the percent relative standard deviation (%RSD)  of
              Relative Response Factor  (RRF) values for each semivolatile target
              compound and DMC over the initial calibration range using Equation
              3  in conjunction with.Equation 4.
              EQ. 3
                                -  (Standard Peviation> ,
                                            Hean
              Where:

               %RSD =       Percent relative standard deviation.

 9.3.4.4      Equation 4  is the general formula for standard deviation for a
              statistically small set of values.

              EQ.  4
                         Staadard Deviation  -
                                                         n-1
 OLC03.2                                  D-20/SV

-------
                                                     Exhibit D Semivolatiles -- Section 9
                                                  Calibration and Standardization (Con't)

              Where:

              Xi, •  and n are as defined in Equation 2.

9.3.5     Technical Acceptance•Criteria for Initial Calibration

9.3.5.1       All initial calibration standards must be  analyzed at  the
              concentration levels described in Section  7.2.3.4.1 and at the
              frequency described in Section 9.3.2 on a  GC/MS system meeting the
              DFTPP technical acceptance criteria.

9.3.5.2       The relative response factor (RRFi)  at  each  calibration
              concentration for each semivolatile target compound and DMC must
              be greater than or equal to the compound's minimum acceptable
              relative response factor listed in Table D-4.

9.3.5.3       The %RSD over the initial calibration range for relative response
              factor for each semivolatile target compound that has  a required
              %RSD must be less than or equal to the %RSD listed in  Table D-4.

9.3.5.4       Up to four compounds may fail the criteria listed in Sections
              9.3.5.2 and 9.3.5.3 and still meet the minimum RRF and %RSD
              requirements.  However, these four compounds must have a minimum
              RRF greater than 0.010 and %RSD less than  or equal to  40.0%.

9.3.5.5       Excluding those ions in the solvent front, no quantitation ion may
              saturate the detector.  Consult the manufacturer's instrument
              manual to determine how saturation is indicated for your
              instrument.

9.3.6     Corrective  Action  for  Initial Calibrations

9.3.6.1.       If-the technical acceptance criteria for initial calibration are
       ..-,      not met, inspect the system for problems.   It may be necessary to
              clean the ion source, change the column, or take other corrective
              actions to achieve the technical acceptance criteria.

9.3.6.2       Initial calibration technical acceptance criteria MUST be met
              before any samples or required blanks are  analyzed. Any samples
              or required blanks analyzed when initial calibration technical
              acceptance criteria have not been met will require re-analysis at
              no additional cost to USEPA.

9.4    Continuing Calibration

9.4.1     Summary  of  Continuing  Calibration

          Prior to the analysis  of samples and required blanks and  after GC/MS
          performance check  technical  acceptance  criteria and  initial
          calibration technical  acceptance criteria have  been met,  each GC/MS
          system must be  routinely checked by analyzing a continuing
          calibration standard to  ensure  that the instrument continues  to meet
          the  instrument  sensitivity and  linearity requirements  of  the  method.
          The  continuing  calibration standard contains  all  the semivolatile
          target compounds,  DMCs,  and internal standards.

9.4.2     Frequency of Continuing  Calibration

9.4.2.1       Each GC/MS used for analysis must be calibrated once every twelve
              (12)  hour time period of operation.   The 12-hour time  period
              begins with the injection of DFTPP.
                                          D-21/SV                                 OLC03.2

-------
Exhibit D Semivolatiles -- Section 9
Calibration and Standardization  (Con't)

9.4.2.2       If time still remains in the 12-hour time period after meeting the
              technical acceptance criteria for the initial calibration, samples
              and blanks may be analyzed.  It is not necessary to analyze a
              continuing calibration standard within this 12-hour time period,
              if the initial calibration standard that is the same concentration
              as the continuing calibration standard meets the continuing
              calibration technical acceptance criteria.  Quantitate all sample
              and blank results against the 20 ng/uL  (80 ng/uL for the seven
              compounds listed in Section 7.2.3.4.1) calibration standard.

9.4.3      Procedure for Continuing Calibration

9.4.3.1       All standard/spiking solutions must be allowed to warm to ambient
              temperature  (approximately 1 hour) before preparation or analysis.

9.4.3.2       Add 10 uL of the internal standard solution  (Section 7.2.3.5) to
              1.0 mL of the continuing calibration standard  (Section 7.2.3.4.3)
              for a concentration of 20 ng/uL for each internal standard
              compound.

9.4.3.3       Analyze the continuing calibration standard by injecting 1.0 uL of
              standard.

9.4.4      Calculations for Continuing Calibration

9.4.4.1       Calculate a RRF for each send/volatile target compound and DMC
              using Equation 1 for the primary characteristic ions found in
              Table D-3.

9.4.4.2       Calculate the percent difference between the mean relative
              response  factor from the most recent  initial calibration and the
              continuing calibration relative response factor for each
              semivolatile target compound and DMC using Equation 5.  For
              internal  standards, use the primary ions listed in Table D-3
              unless interferences are present. . If interferences prevent the
              use of the primary ion for a given internal  standard, use the
              secondary ion(s) listed in Table D-3.

              EQ. 5
               Where:

               	        Mean relative response factor from the most recent
               RRFi =      initial calibration meeting technical acceptance
                          criteria.

               RRFC =      Relative response factor from continuing calibration
                          standard.

 9.4.5     Technical Acceptance Criteria for Continuing Calibration

 9.4.5.1       The continuing calibration standard must be analyzed  at  the 20
               ng/uL (80 ng/uL for the seven compounds listed in 7.2.3.4.1)
               concentration level,  and at the  frequency described in Section
               9.4.2,  on a GC/MS  system meeting the DFTPP and the initial
               calibration technical acceptance criteria.
 OLC03.2                                  D-22/SV

-------
                                                     Exhibit D Semivolatiles -- Section 9
                                                  Calibration and Standardization  (Con't)

9.4.5.2       The relative response factor for each semivolatile target compound
              and DMC must be greater than or equal to the compound's minimum
              acceptable relative response factor listed in Table D-4.

9.4.5.3       The relative response factor percent difference for each
              semivolatile target compound that has a percent difference
              criteria must be within the inclusive range listed in Table D-4.

9.4.5.4       Up to four compounds may fail the requirements listed in Sections
              9.4.5.2 and 9.4.5.3 and still meet the minimum relative response
              factor and percent difference criteria.  However,  these compounds
              must have a minimum relative response factor greater than or equal
              to 0.010 and the percent difference must be within the inclusive
              range of ±40%.

9.4.5.6       Excluding those ions in the solvent front, no quantitation ion may
              saturate the detector.  Consult the manufacturer's instrument
              operating manual to determine how saturation is indicated for your
              instrument.

9.4.6     Corrective Action for Continuing Calibration

9.4.6.1       If the continuing calibration technical acceptance criteria are
              not met, recalibrate the GC/MS instrument according to Section
              9.3.3.  It may be necessary to clean the ion source, change the
              column, or take other corrective actions to achieve the technical
              acceptance criteria.

9.4.6.2       Continuing calibration technical acceptance criteria MUST be met
              before any samples  (including requested MS/MSD) or required blanks
      '-•'•?••      are analyzed.  Any samples or required blanks analyzed when
      •^-      continuing calibration criteria have not been met will require re-
      '-?..-'      analysis at no additional cost to USEPA.
                                          D-23/SV                                 OLC03.2

-------
Exhibit D Semivolatiles -- Section 10
Procedure

10.0   PROCEDURE

10.1   Sample Preparation

10.1.1    This method  is  designed for analysis of water samples that contain
          low concentrations of  the  semivolatile compounds listed  in Exhibit C.
          The majority of the  samples are expected to come from drinking water
          and well/ground water  type sources around Superfund sites.   If, upon
          inspection of a sample, the Contractor suspects that the sample is
          not amenable to this method,  contact Sample Management Office  (SMO).
          SMO will  contact the Region for instructions.

10.1.2    If insufficient sample volume (less than 90 percent of the required
          amount) is received  to perform the analyses, the Contractor  shall
          contact SMO  to  apprise them of the problem.  SMO will contact the
          Region for instructions.   The Region will either require that no
          sample analyses be performed  or will require that  a reduced  volume be
          used  for  the sample  analysis. No other changes in the analyses will
          be permitted.   The Contractor shall document the Region's decision  in
          the Sample Delivery  Group  (SDG) Narrative.

10.1.3    Extraction of Sample

10.1.3.1      Allow the sample to come  to ambient temperature (approximately 1
              hour)  .

10.1.3.2      Continuous  liquid-liquid  extraction is used to extract the
              samples.   Separatory funnel extraction cannot be used.

10.1.3.3      Continuous  Liquid-Liquid  Extraction Without Hydrophobic Membrane

10.1.3.3.1       Follow manufacturer's  instructions for set-up.

10.1.3.3.2       Add methylene chloride to  the bottom of the extractor and  fill
                 it to a  depth of at least  one inch above the bottom sidearm.

10.1.3.3.3       Measure  out a 1 L  sample aliquot in a separate, clean graduated
                 cylinder;  transfer  the aliquot  to the continuous  extractor.
                 Measure  and record  the initial  pH of the sample with  a pH  meter
                 or narrow  range pH  paper.  Adjust the pH to 2.0 with  1:1 H2SO4
                 and record the  final pH.

                 NOTE: With some samples, it may be necessary to place a  layer
                 of glass wool between  the  methylene chloride and  the  water
                  layer in the  extractor to  prevent precipitation of suspended
                  solids  into  the methylene  chloride during extraction.

10.1.3.3.4       Using a  syringe or volumetric pipet, add 1.0 mL of the
                 Deuterated Monitoring  Compound  (DMC) standard  spiking solution
                  (Section 7.2.3.1)  into the sample and mix well.

10.1.3.3.5       Rinse the  graduated cylinder with 50 mL of  methylene  chloride
                  and  transfer  the  rinsate to  the continuous  extractor.  If  the
                  sample container  is empty, rinse  the container with  50  mL  of
                 methylene  chloride and add rinsate  to  the continuous  extractor.

10.1.3.3.6       Add  sufficient  methylene chloride to the continuous  extractor
                  to ensure  proper solvent cycling during operation. Adjust the
                  drip rate  to 5  to 15 mL/minute  (recommended);  optimize  the
                  extraction drip rate.   Extract  for  a minimum of  18 hours.
 OLC03.2                                  D-24/SV

-------
                                                    Exhibit D Semivolatiles -- Section 10
                                                                        Procedure  (Con't)

                 NOTE 1: When a minimum drip rate of 10-15 mLs/min. is
                 maintained throughout the extraction, the extraction time may
                 be reduced to a minimum of twelve hours.  Allow to cool, then
                 detach the distillation flask.  Proceed to Section 10.1.4.

                 NOTE 2: Some continuous liquid-liquid extractors are also
                 capable of concentrating the extract within the extraction set-
                 up.  Follow the manufacturer's instructions for concentration
                 when using this type of extractor.

10.1.3.4      Continuous Liquid-Liquid Extraction With Hydrophobic  Membrane

10.1.3.4.1       Follow the manufacturer's instructions for set-up.-

10.1.3.4.2       Measure out each 1 L sample aliquot in a separate, clean
                 graduated cylinder.  If the sample container is empty, rinse
                 the container with 50 mL of methylene chloride and add the
                 rinsate to the continuous extractor.  If the sample container
                 is not empty, add 50 mL of methylene chloride to the continuous
                 extractor.  Slowly transfer the aliquot to the continuous
                 extractor.  Measure and record the initial pH of the sample
                 with a pH meter or a narrow range pH paper.  Adjust the pH to
                 2.0 with 1:1 H2SO4  and record  the final  pH.

10.1.3.4.3       Using a syringe or volumetric pipet, add 1.0 mL of the DMC
                 standard spiking solution  (Section 7.2.3.1) into the sample and
                 mix well.

10.1.3.4.4       Rinse the graduated cylinder with 50 mL of methylene chloride
      "£••         and transfer the rinsate to the continuous extractor.

10.1.3-.:'4 .5       Add sufficient methylene chloride to the continuous extractor
                 to ensure proper solvent cycling during operation.  Adjust the
      "••         drip rate to 15 mL/minute  (recommended) ; optimize the
      —-'         extraction drip rate.  Extract for a minimum of 6 hours.

      -4"          NOTE 1: Due to the smaller volume of solvent used during the
                 extraction process, some sample matrices  (e.g., oily samples,
                 samples containing a high concentration of surfactants) may
                 create an emulsion which will consume the solvent volume,
                 preventing the efficient extraction of the sample.  When this
                 occurs, add additional solvent to assure efficient extraction
                 of the sample, and extend the extraction time for a minimum of
                 6 hours.  If the sample matrix prevents the free flow of
                 solvent through the membrane, then the non-hydrophobic membrane
                 continuous liquid-liquid type extractor must be used.

                 Allow to cool, then detach the distillation flask.  Proceed to
                 Section 10.1.4.

                 NOTE 2: Some continuous liquid-liquid extractors are also
                 capable of concentrating the extract within the extraction set-
                 up.  Follow the manufacturer's instructions for concentration
                 when using this type of extractor.  Using the hydrophobic
                 membrane type extractor, it may not be necessary to dry the
                 extract with sodium sulfate.

                 NOTE 3: If low DMC recoveries occur, assure 1) the apparatus
                 was properly assembled to prevent leaks,- 2) the drip
                 rate/solvent cycling was optimized; and 3) there was proper
                 cooling for condensation of solvent.
                                          D-25/SV                      -           OLC03.2

-------
Exhibit D Semivolatiles -- Section 10
Procedure (Con't)

                 NOTE 4: Alternate continuous liquid-liquid extractor types that
                 meet the requirements of the SOW may also be used.  If using
                 alternate extractors or design types, follow the manufacturer's
                 instructions for set-up.

10.1.4    Concentrating the  Sample Extract

10.1.4.1      Assemble a Kuderna-Danish (K-D)  concentrator by attaching a 10 mL
              concentrator tube to a 500 mL evaporative flask.   Other
              concentration devices or techniques may be used in place of the
              K-D,  if equivalency is demonstrated for all the semivolatile
              target compounds listed in Exhibit C.

10.1.4.2      Transfer the extract by pouring the extract through a drying
              column containing about 10 cm of anhydrous granular sodium
              sulfate, and collect the extract in a K-D concentrator.  Rinse the
              distilling flask and column with 20 to 30 mL of methylene chloride
              to complete the quantitative transfer.

10.1.4.3      Add one or two clean boiling chips and attach a three-ball Snyder
              column to the evaporative flask.  Pre-wet the Snyder column by
              adding about 1 mL methylene chloride to the top of the column.
              Place the K-D apparatus on a hot water bath  (60°C to 80°C
              recommended) so that the concentrator tube is partially immersed
              in the hot water, and the entire lower rounded surface of the
              flask is bathed with hot vapor.  Adjust the vertical position of
              the apparatus and the water temperature as required to complete
              the concentration in 10 to 15 minutes.  At the proper rate of
              distillation, the balls of the column will actively chatter but
              the chambers will not flood with condensed solvent.  When the
              apparent volume of liquid reaches 1 mL, remove the K-D apparatus
              from the water bath and allow it to drain and cool for at least 10
              minutes.  DO NOT ALLOW THE EVAPORATIVE FLASK TO GO DRY.  Remove
              the Snyder column and rinse the flask and its lower joint into the
              concentrator tube with 1 to 2 mL of methylene chloride.  A 5 mL
              syringe is recommended for this operation.

10.1.4.4      Two different  types of concentration techniques are permitted to
              obtain the final 1.0 mL volume: micro Snyder column and nitrogen
              evaporation techniques.

10.1.4.4.1        Micro Snyder Column Technique

                  Add another one or two clean boiling chips  to the concentrator
                  tube and attach a two-ball micro Snyder column.   Pre-wet the
                  Snyder column by adding about 0.5  mL of methylene chloride to
                  the top of the column.   Place the  K-D apparatus in a hot water
                  bath (60°C to 80°C recommended)  sp that the concentrator tube
                  is partially immersed in the hot water.   Adjust the vertical
                  position of the apparatus and the  water temperature as required
                  to complete the concentration in 5 to 10 minutes.   At the
                  proper rate of distillation the balls of the column will
                  actively chatter but the chambers  will not  flood with condensed
                  solvent.  When the apparent volume of liquid reaches about 0.5
                  mL, remove the K-D apparatus from the water bath and allow it
                  to drain for at least 10 minutes while cooling.   DO NOT LET THE
                  EXTRACT GO DRY. Remove the Snyder column and rinse the
                  evaporative flask and its lower joint into  the concentrator
                  tube with 0.2 mL of methylene chloride.   Adjust the final
                  volume to 1.0 mL with methylene chloride.  Transfer the extract
                  to a PTFE-sealed screw-cap bottle, label the bottle and store
                  at 4°C (±2°C).


 OLC03.2                                  D-26/SV

-------
                                                    Bxhibit D Semivoiatlies  --  Section  10
                                                                        Procedure  (Con't)

10.1.4.4.2       Nitrogen Evaporation Technique  (taken  from ASTM Method D3086)  •

                 Place the concentrator tube  in  a warm  water bath  (30°C to 35°C)
                 and evaporate the solvent volume to just below 1 mL by blowing
                 a gentle stream of clean, dry nitrogen (filtered through a
                 column of activated carbon)  above  the  extract.  Caution: Gas
                 lines from the gas source to the evaporation apparatus must be
                 stainless steel, copper, or  PTFE tubing.  New plastic tubing
                 must not be used between the carbon trap and the sample since
                 it may introduce interferences.  The internal wall of the
                 concentrator tube must be rinsed down  several-times with
                 methylene chloride during the operation and the final volume
                 brought to 1.0 mL with methylene chloride.  During evaporation,
                 the tube solvent level must  be  kept below the water level of
                 the bath.  The extract must  never  be allowed to become dry.
                 Transfer the extract to a PTFE-sealed  screw-cap bottle, label
                 the bottle and store at 4°C  (±2°C).

10.2   Instrument Analysis of Sample

10.2.1    Set up the Gas Chromatograph/ Mass  Spectrometer  (GC/MS)  system per
          the requirements of Section 9.1.  Before  samples or required blanks
          can be analyzed, the instrument must meet the decafluoro-
          triphenylphosphine  (DFTPP), initial calibration, and continuing
          calibration technical acceptance criteria.  All sample, blank
          extracts, and standard/spiking solutions  must be allowed to warm to
          ambient temperature  (approximately  1 hour) before analysis.  All
          sample extracts  [including requested Matrix Spike/Matrix Spike
      ••"•.  Duplicate  (MS/MSD)] and required blanks must  be analyzed under the
       :   same  instrumental conditions as the calibration standards.  ,

10.2.21/x  Add 10.0 uL of the internal standard spiking  solution  (Section
      ?:?•  7.2.3.5) to the 1.0 mL extract.  For sample dilutions, add an
      -%•"  appropriate amount of the  internal  standard spiking solution to
      • '•§}-  maintain a 20 nanograms per microliter (ng/uL) concentration of the
          internal standards in the diluted extract.

          NOTE: An alternate amount of internal  standard solution may be added,
          however the internal standards must be added  to maintain the required
          20 ng/uL of each internal standard  in  the sample extract.

10.2.3    Inject 1.0 uL of sample extract into the  GC/MS, and start data
          acquisition.

10.2.4    Three minutes after all semivolatile target compounds have eluted
          from  the GC, terminate the MS data  acquisition and store data files
          on the data system storage device.  Use appropriate data output
          software to display full range mass spectra and Extracted Ion Current
          Profiles  (EICPs).

10.2.5    Sample Dilutions

          An original undiluted analysis must be made and results  reported for
          all samples.

10.2.5.1      When a sample extract is analyzed that has a semivolatile target
              compound concentration greater than the upper limit of the initial
              calibration range or in which ions from a target  compound saturate
              the detector (excluding the compound peaks in the solvent front),
              the extract must be diluted,  the internal  standard concentration
              must be readjusted,  and the sample extract must be re-analyzed.
              Secondary ion quantitation is only allowed when there are sample


                                          D-27/SV                                  OLC03.2

-------
Exhibit D Semivolatiles -- Section 10
Procedure (Con't)

              interferences with the primary quantitation ion,  not when
              saturation occurs.  If secondary ion quantitation is used,
              calculate a relative response factor using the area response
              (EICP)  from the most intense secondary ion which is free of sample
              interferences,  and document the reasons for the use of the
              secondary ion in the SDG Narrative.

              NOTE:  If the laboratory has evidence or highly suspects, because
              of sample color or other physical property, that a sample may
              contain extremely high concentrations of either target or non-
              target compounds, then SMO shall be immediately contacted.   SMO
              will seek regional recommendations for diluted analysis.

10.2.5.2      Dilute the sample using the following procedure:

10.2.5.2.1        Calculate the sample  dilution necessary to  keep  the
                  semivolatile  target compounds that  required dilution above the
                  mid-point standard  in the  initial calibration range  and so that
                  no  target compound  has ions  which saturate  the detector
                  (excluding  the compound peaks in the  solvent  front).

10.2.5.2.2        Dilute  the  sample extract  quantitatively with methylene
                  chloride.

10.2.5.2.3        Analyze the sample  dilution  per Section 10.2,  including the
                  addition  of internal  standards to maintain  a  20  ng/uL
                  concentration of the  internal standards (Section 10.2.2).
 OLC03.2                                  D-28/SV

-------
                                                    Exhibit D Semivolatiles -- Section 11
                                                           Data Analysis and Calculations

11.0   DATA ANALYSIS AND CALCULATIONS

11.1   Qualitative Identification of Target Compounds

11.1.1    The compounds listed in the Target Compound List  (TCL), Exhibit C,
          shall be  identified by an  analyst competent in the interpretation of
          mass spectra by comparison of the sample mass spectrum to the mass
          spectrum  of a standard of  the suspected compound.  Two criteria must
          be satisfied to verify the identifications.

          •    Elution of the sample analyte within Gas Chromatograph (GC)
               Relative Retention Time (RRT)  unit window established from the
               12-hour calibration standard.

          •    Correspondence of the sample analyte and calibration standard
               component mass spectra.

11.1.2    For establishing'correspondence  of the GC RRT, the sample component
          RRT must  be within ±0.06 RRT units of the RRT of the standard
          component.  For reference, the standard must be run on the same shift
          as the  sample.  For samples analyzed during the same 12-hour time
          period  as the initial calibration standards, compare the analyte
          retention times to those from the 20 nanograms per microliter  (ng/uL)
           (80 ng/uL for the seven compounds listed in Section 7.2.3.4.1)
          calibration standard.  If  coelution of interfering compounds
          prohibits accurate assignment of the sample component RRT from the
          Extracted Ion Current Profile  (EICP) for the primary ion, the RRT
          must be assigned by using  the total ion chromatogram.

11.1.3    For comparison of standard arid sample component mass spectra, mass
          spectra obtained on the Contractor's GC/MS are required.  These
          standard  spectra may be used for identification purposes only if the
          Contractor's GC/MS meets the decafluorotriphenylphosphine  (DFTPP)
          technical acceptance criteria.   These standard spectra may be
          obtained  from the analysis used  to obtain reference RRTs.

11.1.4    The requirements for qualitative verification by comparison of mass
          spectra are as follows:

11.1.4.1      All ions present  in the standard mass spectra at  a relative'
              intensity greater than 10  percent (most abundant  ion in the
              spectrum equals 100 percent)  must be present  in the sample
              spectrum.

11.1.4.2      The relative  intensities of the  major ions  specified in Section
              11.1.4.1 must agree within ±20 percent-between the standard and
              sample spectra.   (Example:  For  an ion with  an abundance of  50
              percent in the  standard spectra,  the corresponding sample  ion
              abundance must  be between  30  and 70  percent.)

11.1.4.3      Ions greater  than 10 percent  in  the  sample  spectrum but not
              present in the  standard spectrum must be considered and accounted
              for by the analyst making  the comparison.   The verification
              process should FAVOR FALSE POSITIVES.   All  compounds meeting the
              identification  criteria must  be  reported with their spectra.  When
              target compounds  are below Contract  Required  Quantitation  Limits
              (CRQLs) but the spectrum meets  the identification criteria,  report
              the concentration with a  "J".  For example, if the CRQL is  5.0
              micrograms per  liter (ug/L)  and  a concentration of 3.0  ug/L is
              calculated,  report the data as  "3.0J".
                                          D-29/SV                                 OLC03.2

-------
Exhibit D Semivolatiles -- Section 11
Data Analysis and Calculations  (Con't)

11.1.5    If a compound cannot  be verified by all of the criteria  in Sections
          11.1.1 through  11.1.4, but  in  the  technical judgment of  the mass
          spectral  interpretation specialist, the identification is correct,
          then the  Contractor shall report that  identification and proceed with
          quantitation.

11.2   Qualitative Identification of Non-Target Compounds

11.2.1    A library search shall be executed for non-target  compounds for the
          purpose of tentative  identification.   The NIST/EPA/NIH  (May 1992
          release or later)  and/or Wiley (1991 release  or  later),  or equivalent
          mass spectral library shall be used as the reference library  (Section
          6.10.5).

11.2.2    Up to  30  non-DMC/non-internal  standard organic compounds of greatest
          apparent  concentration not  listed  in Exhibit  C for the volatiles and
          semivolatiles shall be tentatively identified via  a forward search of
          the NIST/EPA/NIH and/or Wiley  mass spectral  library, or  equivalent
          mass spectral library.  The following  are not to be reported:

          •   Compounds  with responses  less than 10 percent of the internal
               standard (as  determined by inspection of peak areas or heights);

          •   Compounds  which  elute  earlier than 30 seconds before the first
               semivolatile  compound  listed  in Exhibit  C (Semivolatiles)  or
               three minutes after the last  semivolatile compound  listed  in
               Exhibit C  (Semivolatile)  has  eluted; and

           •   Volatile compounds listed in  Exhibit C.  Only after visual
               comparison of sample spectra  to spectra  resulting from the
               library search(es) will the mass  spectral interpretation
               specialist assign a tentative identification.

          NOTE:  Computer generated library search routines must  not  use
          normalization routines that would misrepresent  the library or unknown
           spectra when compared to each other.

 11.2.3    UP to  20 peaks of greatest  apparent  concentration  (as  determined by
           inspection of peak areas or heights)  that are suspected to be
           straight-chain, branched,  or cyclic  alkanes,  alone or  part of an
           alkane series shall be library searched.   When the above alkanes are
           tentatively identified,  the concentration(s)  are to be estimated as
           described  in Section 11.4 and reported in the SDG  Narrative as
           alkanes,  by class (i.e.,  straight chain,  branched, or  cyclic; as a
           series; as applicable).   Alkanes are not counted as part of the 30
           organic compounds described in Section 11.2.2.

 11.2.4     Peaks that are  suspected to be aldol-condensation reaction products
            (e.g., 4-methyl-4-hydroxy-2-pentanone and 4-methyl-3-pentene-2-one)
           shall be  searched, reported, and counted as part of the 30 most
           intense non-target semivolatile compounds,  and qualified with an "A"
         .  flag on Form I  LCSV-TIC.

 11.2.5     Guidelines for  making Tentative Identification

 11.2.5.1     Major ions  in the reference spectrum  (ions greater than 10  percent
              of the most abundant  ion)  should be present  in the sample
              spectrum.

 11.2.5.2     The relative intensities of the major ions should agree within ±20
              percent.   (Example: For an ion with an abundance of  50 percent in
 OLC03.2                                   D-30/SV

-------
                                                    Exhibit D Semivoiatiles -- Section 11
                                                   Data Analysis and Calculations (Con't)

              the  standard spectra,  the corresponding  sample  ion abundance
              should be  between 30 and 70 percent.)

11.2.5.3      Molecular  ions  present in reference  spectrum should be  present  in
              sample spectrum.

11.2.5.4      Ions present in the sample spectrum  but  not  in  the reference
              spectrum should be reviewed for possible background contamination
              or presence of  coeluting compounds.

11.2.5.5      Ions present in the reference spectrum but not  in the sample
              spectrum should be reviewed for possible subtraction from the
              sample spectrum because of background contamination or  coeluting
              compounds.

              NOTE: Data system library reduction  programs can sometimes create
              these discrepancies.

11.2.5.6      Non-target compounds receiving a library search match of 85% or
              higher should be considered a "probable  match".   The compound
              should be  reported unless the mass spectral  interpretation
              specialist feels there is just evidence  not  to  report the compound
              as  identified by the library search  program. The lab should
              include in the  SDG Narrative the justification  for not  reporting a
              compound as listed by the search program.

11.2.5.7      If  the library  search produces more  than one compound at or above
              85%, report the compound with the highest percent match (report
              first compound  if percent match is the same  for two or  more
              compounds), unless the mass spectral interpretation specialist
              feels that the  highest match compound should not be reported, or
              another compound with a lower match  should be reported.  The lab
              should include  in the Sample Delivery Group  (SDG)  Narrative the
              justification for not reporting the  compound with the highest
              spectral match.

11.2.5.8      If  the library  search produces a series  of obvious isomer -
              compounds  with  library search matches greater than 85%  (e.g.,
              tetramethyl naphthalenes), the compound  with the highest library
              search percent  match should be reported  (or  first compound if
              library search  matches are the same). A note should be placed  in
              the  SDG Narrative indicating the exact isomer configuration as
              reported may not be accurate.

11.2.5.9      If  library search matches of less than 85% are  produced and in  the
              technical  judgement of the mass spectral interpretation
              specialist, no  valid tentative identification can be made, the
              compound should be reported as unknown.   The mass spectral
              specialist should give additional classification of the unknown
              compound,  if possible  (e.g., unknown phthalate,  unknown
              hydrocarbon, unknown acid type, unknown  chlorinated compound).  If
              a probable molecular weight can be distinguished,  include it.

11.2.5.10     The  Contractor  shall report pesticide target compounds  listed in
              Exhibit C   (Pesticides) that appear as semivolatile tentatively
              identified compounds.

11.3   Calculations for Target Compounds

11.3.1    Target  compounds identified shall  be  quantitated by the internal
          standard method.   The internal standard used shall  be  the  one
          assigned to that analyte  for quantitation (Table D-2).  The Relative


                                          D-31/SV                       .          OLC03.2

-------
Exhibit D Semivolatiles -- Section 11
Data Analysis and Calculations  (Con't)

          Response  Factor  (RRF)  from the continuing calibration analysis is
          used  to calculate  the concentration in the sample.   For samples
          analyzed  during  the  same  12-hour time period as  the  initial
          calibration standards,  use the RRF values from the mid-point  initial
          calibration standard.

11.3.2    Secondary ion quantitation is allowed only when  there are sample
          interferences with the primary ion.  If secondary ion quantitation is
          performed,  document  the reason in the SDG Narrative.   The area of a
          secondary ion cannot be used for the area of the primary ion  unless a
          relative  response  factor  is calculated using the secondary ion.

          NOTE: Unless otherwise stated, the area response is  from the  EICP of
          the primary quantitation  ion.  The primary quantitation ions  for the
          target compounds,  internal standards, and DMCs are  listed in  Table D-
          3.

11.3.3    Calculate target compound concentrations using Equation 6.

          EQ. 6
           Where:

           AX  =      Area response (EICP)  of the characteristic  ion for the
                      compound to be  measured.

           Ais =      Area response (EICP)  of the characteristic  ion  for the
                      internal standard.   The target  compounds are listed with
                      their associated internal standard in Table D-2.

           Is  '=      Amount of internal  standard injected in nanograms (ng).

           RRF =      Relative-response factor-from the most recent continuing
                      calibration as  determined in Section 9.4.

           V0  =      Volume of water extracted in milliliters (mL).

           Vi  =      Volume of extract injected in microliters (uL).

           Vt  =      Volume of concentrated extract  in microliters (uL).
                      (Vt  = 1000 uL)

           Df   =      Dilution Factor.  The dilution  factor for analysis of
                      water samples for semivolatiles by this method is defined
                      as follows:

           uL  most cone,  extract used to make dilution  + uL clean solvent
                        uL most cone,  extract used to make dilution

           If  no dilution is  performed, Df  = 1.0.

 11.3.4    When a target  compound  concentration  is below the  CRQL,  but  the
           spectra meet the  identification  criteria, report the  concentration
           with a "J".  For  example,  if the CRQL  is 5.0 ug/L and a
           concentration  of  3.0 ug/L is calculated, report as "3.0  J".   Report
           ALL sample concentration  data as UNCORRECTED for blanks.
 OLC03.2                               .    D-32/SV

-------
                                                    Exhibit D Semivolatiles --  Section 11
                                                   Data Analysis and Calculations (Con't)

11.3.5    Calculate the adjusted CRQL for semivolatile compounds using Equation
          7.

          NOTE: If the adjusted CRQL is less than the CRQL listed in Exhibit C
          (Semivolatiles), report the CRQL listed in Exhibit C  (Semivolatiles).

          EQ. 7


                                         Contract                   Contract
                                        Sample Vol                  Injection
                 Adjusted _ Contract, (1000 taL>  --,     Vt>    , Vol  (1 ufl.)
                    CRQL   "   CRQL         v           Contract       V.
                                                         Ext. Vol
                                                         (1000 uL>


          Where:

          Contract  CRQL  =     CRQL values reported in Exhibit C of the SOW.

                       V0  =  Same as EQ. 6.

                       Vi  =  Same as EQ. 6.

                       Vt  =  Same .as EQ. 6.

                       Df  =  Same as EQ. 6.

11.3.6    It is expected that situations will arise where the automated
          quantitation procedures in the GC/MS software provide inappropriate
          quantitations.  This normally occurs when there is compound co-
          elution, baseline  noise, or matrix interferences.  In these
          circumstances, the Contractor must perform a manual quantitation.
          Manual quantitations are performed by integrating the area of the
          quantitation ion of the compound.  This integration shall only
          include the area attributable to the specific TCL, DMC, or internal
          standard compound.  The area integrated shall not include baseline
          background noise.  The area integrated shall not extend past the
          point where the sides of the peak intersect with the baseline noise.
          Manual integration is not to be used solely to meet QC criteria, nor
          is it to be used as a substitute for corrective action on the
          chromatographic system.  Any instance of manual integration must be
          documented in the  SDG Narrative.

11.3.7    In all instances where the data system report has been edited, or
          where manual integration or quantitation has been performed, the
          GC/MS operator must identify such edits or manual procedures by
          initialing and dating the changes made to the report and shall
          include the integration scan range.  In addition, a hardcopy printout
          of the EICP of the quantitation ion displaying the manual integration
          shall be included  in the raw data.  This applies to all compounds
          listed in Exhibit  C (Semivolatiles), internal standards, and DMCs.

11.3.8    The requirements listed in 11.3.4 and 11.3.5 apply to all standards,
          samples  [including requested Matrix Spike/Matrix Spike Duplicate
           (MS/MSD)], and blanks.

11.3.9    Internal Standard  Responses and Retention Times

          Internal standard  responses and retention times in all samples and
          blanks must be evaluated during or immediately after data
          acquisition.  Compare the sample/blank internal standard responses


                                         D-33/SV                                  OLC03.2

-------
Exhibit D Semivolatiles -- Section 11
Data Analysis and Calculations  (Con't)

          and retention times to  the  continuing calibration internal standard
          responses and retention times.   For samples and blanks analyzed
          during the  same  12-hour time period as the initial calibration
          standards,  compare the  internal  standard responses and retention
          times against the 20  ng/uL  (80 ng/uL for the seven compounds  listed
          in Section  7.2.3.4.1) calibration  standard.  The EICP of the  internal
          standards 'must be monitored and  evaluated for each sample and blank.

11 . 4   Calculations for Non-Target Compounds

       Equation 6 is used for calculating the concentrations of the non-target
       compounds.  Total area. counts  (or peak heights)  from the Reconstructed
       Ion Chromatograms (RICs)  are to be used for both the non-target compound
       to be 'measured  (A*)  and the internal standard  (Ais) .  Associate the
       nearest internal standard  free of interferences with the non-target
       compound to be measured.  A RRF of one (1)  is to be assumed.  The value
       from this quantitation shall be qualified as "J" (estimated due to lack
       of a compound-specific relative response factor) ,  and "N" (presumptive
       evidence of presence) , indicating the qualitative and quantitative
       uncertainties associated with  this non-target compound.  This estimated
       concentration must be calculated for all tentatively identified
       compounds as well as those identified as unknowns.

11.5   Calculations for Deuterated Monitoring Compounds (DMCs)

11.5.1     Calculate the  concentration of  the DMCs using the  same equation as
           used for the target  compounds.

11.5.2     Calculate the  DMC percent recovery in all  samples  and blanks  using
           Equation 8. Determine  if recovery is within limits  (Table D-5)  and
           report on appropriate form.

           EQ.  8

                              Percent fttccvtry - — « 100
           Where:

           Qd =   Concentration or amount determined by analysis.

           Qa =   Concentration or amount added to sample/blank.


 11 . 6   Technical Acceptance  Criteria for Sample Analysis

 11.6.1    The sample must be analyzed on a GC/MS system meeting the GC/MS
           performance check, initial calibration, and continuing calibration
           technical acceptance criteria.

 11.6.2    The sample must be extracted and analyzed within the contract holding
           times.

 11.6.3    The sample must have an associated method blank meeting the blank
           technical acceptance criteria.

 11.6.4    The percent recovery for the DMCs in the sample must be within the
           acceptance windows listed in Table D-5. Up to four DMCs per sample
           may fail to meet the recovery limits listed in Table D-5.
 OLC03.2                                  D-34/SV

-------
                                                    Exhibit D Semivolatiles -- Section 11
                                                   Data Analysis and Calculations  (Con't)

          NOTE: The DMC recovery requirements do not apply to a sample that has
          been diluted.

11.6.5    The instrumental area response  (EICP area) for each of the internal
          standards in the sample must be within the inclusive range of -50
          percent and +100 percent of its response in the most recent
          continuing calibration standard analysis.

11.6.6    The retention time shift for each of the internal standards in the
          sample must be within ±0.33 minutes  (20.0 seconds) of its retention
          time in the most recent continuing calibration standard analysis.

11.6.7    The RRT of each DMC  in the sample must be within ±0.06 RRT units of
          its relative retention time in the most recent continuing calibration
          standard analysis.

11.6.8    Excluding those ions in the solvent front, no ion may saturate the
          detector.  No target compound concentration may exceed the upper
          limit of the initial calibration range unless a more dilute aliquot
          of the sample extract is also analyzed according to the procedures in
          Section 10.2.5.

11.7   Corrective Action for Sample Analysis

11.7.1    If the sample technical acceptance criteria for the DMCs and internal
   -..'•v~f'..  standards are not met, check calculations, DMC and internal standard
          solutions, and instrument performance.  It may be necessary to
          recalibrate the instrument or take other corrective action procedures
  ••-.. •  ::'.••  to meet the DMC and  internal standard technical acceptance criteria.

11.7.2' •   If the Contractor needs to analyze more than one  (1) sample dilution
          other than the original analysis to have all the target compounds
          within the initial calibration range and to have no ions saturating
          the detector  (excluding.the peaks in the solvent front), contact
      ,   Sample Management Office  (SMO).  SMO will contact the Region for
          instructions.

11.7.3    Corrective actions for failure to meet instrument performance checks,
          initial calibration, continuing calibration and method blanks must be
          completed before the analysis of samples.

11.7.4    Sample analysis technical acceptance criteria MUST be met before data
          is reported.  Samples contaminated from laboratory sources or
          associated with a contaminated method blank --or any samples
          analyzed not meeting the technical acceptance criteria -- will
          require re-extraction and/or re-analysis at no additional cost to
          USEPA.

11.7.5    Sample reruns performed as a result of suspected matrix interferences
          beyond the scope of  the method will be reviewed on a case-by-case
          basis for payment purposes by the USEPA Contract Laboratory Program
          Project Officer  (CLP PO).  Send a copy of the SDG Narrative
           (including your contract number), a description of the situation, and
          the  requested action to the CLP PO.
                                          D-35/SV                                 OLC03.2

-------
Exhibit D Semivolatiles -- Section 12
Quality Control

12.0   QUALITY  CONTROL

12.1   Method Blank

12.1.1    Summary of Method Blank

          A method blank  is 1.0  liter  (L)  of  reagent  water  carried through the
          entire analytical scheme.  The purpose  of  a  method blank  is  to
          determine the levels of  contamination  associated  with  the processing
          and  analysis of samples.

12.1.2    Frequency of Method Blank

12.1.2.1      A method blank must be extracted each time samples are extracted.
              The number of samples extracted with each method blank shall not
              exceed 20 field samples  (excluding Matrix Spike/Matrix Spike
              Duplicate(s) MS/MSD, if required, and Performance Evaluation  (PE)
              samples).

12.1.2.2      Each method blank must be analyzed on each Gas Chromatograph/Mass
              Spectrometer (GC/MS) system used to analyze the samples prepared
              with the method blank.

12.1.3    Procedure  for Method Blank

          -Measure  out  1.0 L,of•reagent-water for. each, method  blank aliquot.
          Extract,  concentrate and analyze the method blank at the same time as
          the  samples  associated with the blank according to  Section 10.

12.1.4   Calculations for Method Blank

          Perform data analysis  and calculations according to Section 11.

12.1.5   Technical  Acceptance  Criteria For Method Blank

12.1.5.1      All blanks  must be analyzed  at  the/frequency  described  in Section
             •12.1.2 on a GC/MS  system meeting, the GC/MS-performance  check,
               initial calibration, and continuing calibration technical
               acceptance  criteria.

12.1.5.2       The percent recovery for each of the Deuterated Monitoring
               Compounds  (DMCs) in the  blank must be within  the acceptance
               windows listed in  Table  D-5.

12.1.5.3       The area response  for  each of the  internal  standards in the blank
               must be within the inclusive range of -50 percent  and +100  percent
               of its response in the most  recent continuing calibration standard
               analysis.

12.1.5.4       The retention time shift for each  of  the internal  standards in the
               blank must  be within ±0.33 minutes  (20.0 seconds)  of its  retention
               time in the most recent  continuing calibration standard analysis.

12.1.5.5       The Relative Retention Time  (RRT)  of  each DMC in the blank  must be
               within ±0.06 RRT units of its relative  retention time in  the most
               recent continuing  calibration standard  analysis.

 12.1.5.6       The concentration  of all target compounds  (except  the phthalate
               esters listed  in Exhibit C)  in  the blanks must be  less  than the
               Contract Required  Quantitation  Limit  (CRQL) for each target
               compound.   A method blank for semivolatile  analysis  must  contain
               less than five times  (5X) the CRQL of the phthalate  esters  listed


OLG03.2                                   D-36/SV

-------
                                                    Exhibit D Semivolatiles -- Section 12
                                                                  Quality Control  (Con't)

              in Exhibit C.   The concentration of non-target compounds in the
              blanks must not exceed 10 micrograms per liter (ug/L).

12.1.6    Corrective Action  for Method.Blank

12.1.6.1      If a Contractor's blank does not meet the technical acceptance
              criteria  for method blanks the Contractor shall consider the
              analytical system to be out of control.

12.1.6.2      It is the Contractor's responsibility to ensure that method
              interferences caused by contaminants in solvents,  reagents,
              glassware, and sample storage and processing hardware that lead to
              discrete  artifacts and/or elevated baselines in gas chromatograms
              be eliminated.

12.1.6.3      If contamination is a problem, the source of the contamination
              must be investigated and appropriate corrective measures MUST be
              taken and documented before further sample analysis proceeds.
              Further,  all samples (including requested MS/MSD and PE samples)
              processed with a method blank that does not meet the blank
              technical acceptance criteria (i.e., contaminated)  will require
              re-extraction and re-analysis at no additional cost to USEPA.

12.1.6.4      If DMC recoveries in the method blank do not meet the technical
              acceptance criteria (Section 12.1.5.2),  first re-analyze the
  ...  :^i;       method blank.   If the DMC recoveries do not meet the technical
      ~        acceptance criteria after re-analysis,  then all samples
      4'       (including requested MS/MSD and PE samples) associated with that
              method blank must be re-extracted and re-analyzed at.no additional
              cost to USEPA.

12.1.6.5      If the method blank fails to meet a technical acceptance criteria
              other than Sections 12.1.5.6 and 12.1.5.2, then the problem is an
      -•£»       instrument problem.   Correct the instrument problem, recalibrate
              the instrument (if necessary), and re-analyze the method blank.
      ~'~      Sample analysis cannot proceed until the method blank meets these
  •*~ ;:n-       technical acceptance requirements.

12.2  Matrix Spike/Matrix Spike Duplicate  (MS/MSD)

12.2.1    Summary of MS/MSD

          In  order to  evaluate  the effects of  the sample matrix  on the methods
          used  for semivolatile analyses,  USEPA has prescribed a mixture of
          semivolatile target compounds  to be  spiked  into two aliquots of a
          sample  and analyzed in  accordance with the  appropriate method upon
          request.

12.2.2    Frequency of MS/MSD Analyses

12.2.2.1      A matrix  spike and matrix spike duplicate shall only be analyzed
              if requested by the Region [(through the Sample Management Office
              (SMO)] or specified on the Traffic Report  (TR).  If requested,  a
              matrix spike and matrix spike duplicate must be extracted and
              analyzed  for every 20 field samples in a Sample Delivery Group
              (SDG), or each SDG,  whichever is most frequent.

12.2.2.2      As part of USEPA's QA/QC program, water rinsate samples and/or
              field/trip blanks (field QC)  may be delivered to the laboratory
              for analysis.   The Contractor shall not perform MS/MSD analysis on
              any of the field QC samples.
                                          D-37/SV                                 OLC03.2

-------
Exhibit D Semivolatiles -.- Section 12
Quality Control (Con't)

12.2.2.3      If the USEPA Region requesting MS/MSD designates a sample to be
              used as an MS/MSD, then that sample must be used.  If there is
              insufficient sample remaining to perform an MS/MSD,  then the
              Contractor shall choose another sample on which to perform an
              MS/MSD analysis.  At the time the selection is made,  the
              Contractor shall notify the Region (through SMO) that insufficient
              sample was received and identify the USEPA sample selected for the
              MS/MSD analysis.  The rationale for the choice of another sample
              other than the one designated by USEPA shall be documented in the
              SDG Narrative.

12.2.2.4      If there is insufficient sample remaining in any of the samples in
              an SDG to perform the requested MS/MSD, then the Contractor shall
              immediately contact SMO to inform them of the problem.  SMO will
              contact the Region for instructions.  The Region will either
              approve that no MS/MSD is required, or require that a reduced
              sample aliquot be used for the MS/MSD analysis.  SMO will notify
              the Contractor of the Region's decision.  The Contractor shall
              document the decision in the SDG Narrative.

12.2.2.5      If it appears that the Region has requested MS/MSD analysis at a
              greater frequency than specified in Section 12.2.2.1, then the
              Contractor shall contact SMO.  SMO will contact the Region to
              determine which samples should have MS/MSD performed on them.  SMO
              will notify the Contractor of the Region's decision.  The
              Contractor shall document-the decision,in the SDG Narrative.  If
              this procedure  is not followed, the Contractor will not be paid
              for MS/MSD analysis performed at a greater frequency than required
              by the contract.

12.2.2.6      When a Contractor receives only PE samples, no MS/MSD shall be
              performed within  that SDG.

12.2.2.7      When a Contractor receives a performance evaluation  (PE) sample as
              part of a larger  SDG, a sample other than  the PE sample must be
              chosen for the  requested MS/MSD analysis when the Region did not
              designate a sample to be used for this purpose.

12.2.3    Procedure for Preparing MS/MSD

12.2.3.1      Measure out two additional 1 L aliquots of the  sample chosen for
              spiking in two  continuous extractors.  Add 1.0 mL of DMC spiking
              solution  (Section 7.2.3) and 1.0 mL of the matrix spiking solution
               (Section 7.2.3.2) to each aliquot.  Extract, concentrate, and
              analyze the MS/MSD according to the procedures  described in
              Section 10.

12.2.4    Dilution of MS/MSD

           MS/MSD samples must be analyzed at the same concentration as the most
           concentrated extract for which the original sample results will be
           reported.  For example, if the original sample is to be reported at a
           1:1 dilution and a 1:10 dilution, then analyze and report the MS/MSD
           at a 1:1 dilution only.  However, if the original sample is to be
           reported at a 1:10 dilution and a 1:100 dilution, then the MS/MSD
           must be analyzed and reported at a 1:10 dilution only.  Do not
           further dilute the MS/MSD samples to get either spiked or non-spiked
           analytes within calibration range.  Dilution of the sample must be
           performed in accordance to the conditions in Section 10.2.5.
 OLC03.2          .                        D-38/SV

-------
                                                    Exhibit D Semivolatiles -- Section 12
                                                                  Quality Control (Con't)
12.2.5    Calculations for MS/MSD
12.2.5.1      Calculate the recovery of each matrix  spike .compound in the matrix
              spike and matrix spike duplicate  samples and  report on the
              appropriate form.   Calculate  the  concentrations of the matrix
              spike compounds using the same equation as used for target
              compounds (Equation 6).  Calculate  the recovery of each matrix
              spike compound as follows:

              EQ.  9
                                                       - SR
                               Spike Recovery •  ±±	_ * 100
              Where:

              SSR  =      Spike sample result.

              SR  =       Sample result.

              SA  =       Spike added.
12.2.5.2      Calculate the Relative Percent Difference  (RPD)  of  the recoveries
              of each compound in the matrix spike  and matrix  spike duplicate as
              follows:
              EQ.  10
                              RPD -   I*8*"  "*'   » 100
                                     = (HSR + KSDR>
              Where:

              MSR   =       Matrix spike recovery.

              MSDR  =       Matrix spike duplicate  recovery.


12.2.6    Technical Acceptance Criteria for MS/MSD

12.2.6.1      If requested,  all  MS/MSD must be  prepared and analyzed  at the
              frequency described in Section  12.2.2.  All  MS/MSD must be
              analyzed on a  GC/MS system meeting decafluorotriphenylphosphine
              (DFTPP),  initial and continuing calibration  technical acceptance
              criteria,  and  the  method blank  technical  acceptance criteria.

12.2.6.2      The MS/MSD must have an associated method blank meeting the blank
              technical acceptance criteria.

12.2 .6.3      The MS/MSD must be extracted and  analyzed within the contract
              holding time.

12.2.6.4      The retention  time shift for each of  the  internal standards must
              be within ±0.33 minutes (20 seconds)  between the MS/MSD sample and
              the most recent continuing calibration  standard.

12.2.6.5      The relative retention time for the DMCs  must be within ±0.06 RRT
              units of its standard retention time  in the  continuing  calibration
              standard.


                                          D-39/SV                                'OLC03.2

-------
Exhibit D Semivolatiles -- Section 12
Quality Control (Con't)

12.2.6.6      The  limits for matrix spike compound recovery and RPD are given in
              Table D-6.  As these limits are only advisory,  no further action
              by the laboratory is required;  however,  frequent failures to meet
              the  limits for recovery or RPD warrant investigation by the
              laboratory,  and may result in questions from USEPA.

12 .2.6.7      Corrective Action for Matrix Spike/Matrix Spike Duplicate (MS/MSD)

              Any MS/MSD which fails to meet the technical acceptance criteria
              in Sections 12.2.6.1 through 12.2.6.5 must be re-analyzed at no
              additional cost to USEPA.

12 .3   Method Detection Limit (MDL)  Determination

12.3.1    Before  any field samples  are analyzed under  this contract,  the  MDL
          for each semivolatile target compound shall  be  determined for each
          sample  extraction procedure and on one of the instruments to be used
          for sample analysis.   The MDLs must be verified annually thereafter
           (see Section 12.3.2 for MDL verification procedures), until the
          contract expires or is terminated  or  after major  instrument
          maintenance.  Major instrument maintenance includes,  but is not
          limited to cleaning or replacement of the mass  spectrometer source,
          mass filters (e.g., quadrupole,  ion trap,  etc.),  or electron
          multiplier (or  similar device) ,  or replacement  of  gas chromatographic
          column.

12.3.2    To determine the MDLs, the Contractor shall  run an MDL  study
          following the procedures  specified in 40.CFR Part  136.   The
          Contractor is only -required to analyze the MDL  samples  on one
          instrument used for field sample analyses.   MDL verification only is
          then required on all other instruments used  for field sample analysis
          and at  the frequency specified in Section 12.3.1.   MDL  verification
          is achieved by analyzing a single  reagent water blank spiked with  .
          each target compound at a concentration equal to two times the
          analytical determined MDL.  The resulting mass  spectra  of each target
          compound must meet the qualitative identification criteria outlined
          in Sections 11.1.1 through 11.1.4.3.

12.3.3    The determined concentration of the MDL must be less than the CRQL.

12.3.4    All documentation for the MDL studies shall  be maintained at the
           laboratory and provided to USEPA upon written request.
 OLC03.2                                  D-40/SV

-------
                                                Exhibit  D Semivolatiles  --  Sections 13-16
                                                                       Method  Performance

13.0   METHOD PERFORMANCE

       Not Applicable.

14 . 0   POLLUTION PREVENTION

14.1   Pollution prevention encompasses any technique that reduces or
       eliminates the quantity or toxicity of waste at the point of generation.
       Numerous opportunities for pollution prevention exist in laboratory
       operation.  USEPA has established a preferred hierarchy of environmental
       management techniques that places pollution prevention as the management
       option of first choice.  Whenever feasible, laboratory personnel should
       use pollution prevention techniques to address their waste generation.
       When wastes cannot be feasibly reduced at the source, USEPA recommends
       recycling as the next best option.

.14.2   For information about-pollution prevention that may be applicable to
       laboratories and research institutions, consult Less is Better:
       Laboratory Chemical Management for Waste Reduction, available from the
       American Chemical .Society's Department of Government Relations and
       Science Policy, 1155 16th Street N.W., Washington D.C.  20036,  (202)
       872-4477.

15.0   WASTE MANAGEMENT

       USEPA requires that laboratory waste management practices be conducted
      ^consistently with all applicable rules and regulations.  USEPA urges
       •laboratories to protect the air, water, and land  by minimizing and
       'controlling all releases from hoods and bench operations, complying with
      /the letter and spirit of any sewer discharge permits and regulations,
       :and by complying with all solid and hazardous waste regulations,
       particularly the hazardous waste identification rules and land disposal
       restrictions.  For further information on waste management consult The
       ^Waste Management Manual for Laboratory Personnel,  available from the
       American Chemical Society at the address listed in Section 14.2.

16 . 0  ..'REFERENCES

       Not Applicable.
                                          D-41/SV                                 OLC03.2

-------
Exhibit D Semivolatiles -- Section 17
Tables/Diagrams/Flowcharts

17.0   TABLES/DIAGRAMS/FLOWCHARTS
                                        Table D-l
                        DFTPP KEY IONS AND ION ABUNDANCE CRITERIA

            Mass	Ion Abundance Criteria

             51       30.0-80.0 percent of mass 198
             68       Less than 2.0 percent of mass 69
             69       Present
             70       Less than 2.0 percent of mass 69
             127      25.0 - 75.0 percent of mass 198
             197      Less than 1.0 percent of mass 198
             198      Base peak, 100 percent relative abundance  (see note)
             199      5.0 - 9.0 percent of mass 198
             275      10.0 - 30.0 percent of mass 198
             365      Greater than 0.75 percent of mass 198
             441      Present but less than mass 443
             442      40.0 - 110.0 percent of mass 198
             443      15.0 - 24.0 percent of mass 442

       NOTE: All ion abundances MUST be normalized to m/z 198, the nominal base
       peak, even though  the ion abundances of m/z 442 may be up to 110 percent
       that  of  m/z 198.
 OLC03.2                                  D-42/SV

-------
                                                             Exhibit D Semivolatiles -- Section 17
                                                                Tables/Diagrams/Flowcharts (Con't)
                                                 Table D-2
                            Send/volatile Internal Standards With Corresponding
                   Target and Deuterated Monitoring Compounds Assigned, for Quantitation
        l,4-Dichlorobenzene-d4
         Naphthalene-da
        Acenaphthene - d10
Ben z a 1 dehyde
Phenol
bis(2-Chloroethyl)ether
2-Chlorophenol
2-Methylphenol
2,2'-oxybis-(1-Chloro-propane)
Ace tophenone
4-Methylphenol
N-Nitroso-Di-n-propylamine
Hexachloroethane
Phenol-ds (DMC)
Bis(2-chloroethyl)ether-d8  (DMC)
2-Chlorophenol-d4 (DMC)
4-Methylphenol-ds (DMC)
Nitrobenzene
Isophorone
2-Nitrophenol
2,4-Dimethylphenol
bis(2-Chloroethoxy)methane
2,4-Dichlorophenol
4 -Chioroani1ine
Hexachlorobutadiene
Caprolactam
4-Chloro-3-methylphenol
2-Methylnaphthalene
Naphthalene
Nitrobenzene-ds (DMC)
2 -Nitrophenol -d4 (DMC)
2,4-Dichlorophenol-d3 (DMC)
4-Chloroaniline-d, (DMC)
Hexachlorocyclopentadiene
2,4,6-Trichlorophenol
2,4,5-Trichlorophenol
1,1'-Biphenyl
2 - Chi or onapht ha 1 ene
2-Nitroaniline
Dimethylphthalate
Acenaphthylene
3-Nitroaniline
Acenaphthene
2,4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2,4-Dinitrotoluene
2,6-Dinitrotoluene
1,2,4,5-Tetrachlorobenzene
Diethylphthalate
4-Chlorophenyl-phenylether
Fluorene
4-Nitroaniline
Acenaphthylene-d8 (DMC)
4-Nitrophenol-d« (DMC)
Dimethylphthalate-d6 (DMC)
Fluorene-d10 (DMC)
                                                  D-43/SV
                                                 OLC03.2

-------
        Exhibit D Semivolatiles  -- Section  17
        Tables/Diagrams/Flowcharts  (Con't)

                                             Table  D-2  {Con't)

                            Semivolatile Internal Standards With Corresponding
                   Target and Deuterated Monitoring Compounds Assigned for Quantitation
             Phenanthrene-d.
          Chrysene-d12
        Perylene-d12
4,6-Dinitro-2-methylphenol
N-nitrosodiphenylamine
4-Bromophenyl-phenylether
Hexachlorobenzene
Atrazine
Pentachlorophenol
Phenanthrene
Anthracene
Di-n-butylphthalate
Fluoranthene
4,6-Dinitro-2-methylphenol-d2  (DMC)
Anthracene-dio  (DMC)
Pyrene
Butylbenzylphthalate
3,3'-Dichlorobenzidine
Benzo(a)anthracene
bis(2-Ethylhexyl)phthalate
Chrysene
Pyrene-d10  (DMC)
Di-n-octylphthalate
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo(a)pyrene
Indeno(1,2,3-cd)pyrene
Dibenzo(a,h)anthracene
Benzo(g,h,i)perylene
Benzo(a)pyrene-d12 (DMC)
         OLC03.2
     D-44/SV

-------
                                   Exhibit D  Semivolatiles  --  Section  17
                                     Tables/Diagrams/Flowcharts  (Con't)
                       Table  D-3

Characteristic Ions for Semivolatile Organic Compounds
Target Compounds
Ben z a 1 dehyde
Phenol
bis (2-Chloroethyl) ether
2 -Chlorophenol
2 -Methylphenol
2,2' -oxybis (1-Chloropropane)
Acetophenone
4 -Methylphenol
N-nitroso-di-n-propylamine
Hexachloroethane
Nitrobenzene
Isophorone
2-Nitrophenol
2 , 4 -Dimethylphenol
bis ( -2 -Chloroethoxy) methane
2 , 4 -Dichlorophenol
Naphthalene
4 -Chloroaniline
Hexachlorobutadiene
Caprolactam
4 -Chloro- 3 -methylphenol
2 -Methylnaphthalene
Hexachlorocyclopentadiene
2 , 4 , 6-Trichlorophenol
2 , 4 , 5-Trichlorophenol
1,1' -Biphenyl
2 - Chloronaphthalene
2-Nitroaniline
Dimethyl phthalate
Acenaphthylene
3 -Nitroaniline
Acenaphthene
2 , 4 -Dinitrophenol
4-Nitrophenol
Dibenzofuran
2 , 4 -Dinitrotoluene
2 , 6 -Dinitrotoluene
Diethylphthalate
1,2,4,5 -Tetrachlorobenzene
4 -Chlorophenyl -phenylether
Fluorene
Primary Ion
77
94
93
128
108
45
105
108
70
117
77
82
139
107
93
162
128
127
225
113
107
142
237
196
196
154
162
65
163
152
138
153
184
109
168
165
165
149
216
204
166
Secondary Ion(s)
105, 106
65, 66
63, 95
64, 130
107
77, 79
77, 51
107
42, 101, 130
201, 199
123, 65
95, 138
65, 109
121, 122
95, 123
164, 98
129, 127
129
223, 227
55, 56
144, 142
141
235, 272
198, 200
198, 200
153, 76
164, 127
92, 138
194, 164
151, 153
108, 92
152, 154
63, 154
139, 65
139
63, 182
89, 121
177, 150
214, 179, 108, 143, 218
206, 141
165, 167
                        D-45/SV
OLC03.2

-------
Exhibit D Semivolatiles -- Section 17
Tables/Diagrams/Flowcharts  (Con't)
                                    Table D-3  (Con't)

                  Characteristic Ions for Semivolatile Organic Compounds
Target Compounds
4-Nitroaniline
4 , 6 -Dinitro-2 -methylphenol
N-Nitrosodiphenylamine
4 -Bromophenyl -phenylether
Hexachlorobenzene
Atrazine
Pentachlorophenol
Phenanthrene
Anthracene
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Benzo (a) anthracene
bis (2-Ethylhexyl)phthalate
Chrysene
Di-n-octyl phthalate
Benzo (b) f luoranthene
Benzo (k) f luoranthene
Benzo (a) pyrene
Indeno ( 1 , 2 , 3 - cd ) pyrene
Dibenz (a, h) anthracene
Benzo (g , h , i ) perylene
Primary Ion
138
198
169
248
284
200
266
178
178
149
202
202
149
252
228
149
228
149
252
252
252
276
278
276
Secondary Ion(s)
92,
182
168,
250,
142,
173,
264,
179,
179,
150,
101,
101,
91,
254,
229,
167,
226,

253,
253,
253,
138,
139,
138,
108
, 77
167
141
249
215
268
176
176
104
100
100
206
126
226
279
229
-
125
125
125
227
279
277
 Deuterated Monitoring Compounds
 Phenol-ds                                     99                                   71, 42
 bis-(2-Chloroethyl)ether-d8                   67                                   99, 69
 2-Chlorophenol-cU                             132                              134,  68, 66
 4-Methylphenol-ds                             113                                  115, 54
 Nitrobenzene-d5                               128                                   82, 54
 2-Nitrophenol-d4                              143                               69,  41, 42
 2,4-Dichlorophenol-d3                         165                                 167, 101
 4-Chloroaniline-d4                            131                         .         133, 69
 Dimethylphthalate-d6.                          166                                       78
 Acenaphthylene-d8                             160                                  80, 158
 4-Nitrophenol-d,                              143                              113,  41, 42
 Fluorene-d10                                   176                              174,  87, 86
 4,6-Dinitro-2-methylphenol-d2                 200                                  170, 52
 Anthracene-d10                                188                                   94, 80
 Pyrene-d10                                    212                                 106, 104
 Benzo(a)pyrene-d12                            264                                 132, 118

 OLC03.2                                   D-46/SV

-------
                                                   Exhibit  D  Semivolatiles  --  Section 17
                                                      Tables/Diagrams/Flowcharts (Con't)
                                    Table D-3  (Con't)

                 Characteristic Ions for Semivolatile Organic Compounds
Internal Standard Compounds               Primary Ion                   Secondary Ion(s)
l,4-Dichlorobenzene-d4                       152            .                         115
Naphthalene-d8                               136  .                                    68
Acenapthene-d10                              164                                162,  160
Phenanthrene-d10                             188                                  94,  80
Chrysene-d12                                 240                                120,  236
Perylene-d12	   	264	260,  265
                                         D-47/SV                                  OLC03.2

-------
Exhibit D Semivolatiles - Section 17
Tables/Diagrams/Flowcharts  (Con't)
                                        Table D-4
              Acceptance Criteria for  Initial  and Continuing Calibration of
            Semivolatile Target  Compounds and Deuterated Monitoring Compounds
Semivolatile Compounds
Benzaldehyde
Phenol
bis- (2-Chloroethyl) ether
2 -Chlorophenol
2 -Methylphenol
2,2' -oxybis- (1-Chloropropane)
Acetophenone
4 -Methylphenol
N-Nitroso-di-n-propylamine
Hexachloroethane
Nitrobenzene
Isophorone
2 -Nitrophenol
2 , 4 -Dimethylphenol
bis- (2-Chloroethoxy) methane
2 , 4 -Dichlorophenol
Naphthalene
4 -Chloroaniline
Hexachlorobutadiene
Caprolactam
4 -Chloro- 3 -Methylphenol
2 -Methylnaphthalene
Hexachlorocyclopentadiene
2,4,6 -Trichlorophenol
2,4,5 -Trichlorophenol
1,1' -Biphenyl
2 -Chloronaphthalene
2 -Nitroaniline
Dimethylphthalate
2 , 6-Dinitrotoluene
Acenaphthylene
3 -Nitroaniline
Acenaphthene
2 , 4 -Dinitrophenol
4 -Nitrophenol
Dibenzofuran
2 , 4 -Dinitrotoluene
Diethylphthalate
1,2,4, 5-Tetrachlorobenzene
4 -Chlorophenyl -phenylether
Fluorene
4 -Nitroaniline
4 , 6 -Dinitro-2 -Methylphenol
4 -Bromophenyl -phenylether
N-Nitrosodiphenylamine
Hexachlorobenzene
Atrazine
Pentachlorophenol
Phenanthrene
Minimum RRF
0.010
0.800
0.700
0.800
0.700
0.010
0.010
0.600
0.500
0.300
0.200
0.400
0.100
0.200
0.300
0.200
0.700
0.010
0.010
0.010
0.200
0.400
0.010
0.200
0.200
0.010
0.800
0.010
0.010
0.200
0.900
0.010
0.900
0.010
0.010
0.800
0.200 .
0.010
0.010
0.400
0.900
0.010
0.010
0.100
0.010
0.100
0.010
0.050
0.700
Maximum %RSD
none
20.5
20.5
20.5
20.5
none
none
20.5
20.5
20.5
20.5
20.5
30.0
30.0
20.5
20.5
20.5
none
none
none
20.5
20.5
none
20.5
20.5
none
20.5
none
none
20.5
20.5
none
20.5
none
none
20.5
30.0
none
none
20.5
20.5
none
none
20.5
none
20.5
none
20.5
20.5
Maximum %Diff
none
±25.0
±25.0
±25.0
±25.0
none
none
±25.0
±25.0
±25.0
±25.0
±25.0
' ±30.0
±30.0
±25.0
±25.0
±25.0
none
none
none
±25.0
±25.0
none
±25.0
±25.0
none
±25.0
none
none
±25.0
±25.0
none
±25.0
none
none
±25.0
±30.0
none
none
±25.0
±25.0
none
none
±25.0
none
±25.0
none
±25.0
±25.0
 OLC03.2
D-48/SV

-------
                                        Exhibit D Semivolatiles -- Section 17
                                           Tables/Diagrams/Flowcharts  (Con't)

                        Table D-4  (Con't)

 Acceptance  Criteria  for  Initial  and Continuing  Calibration of
Semivolatile Target Compounds and Deuterated Monitoring Compounds
Semivolatile Compounds
Anthracene
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Benzo (a) anthracene
Chrysene
bis- (2-Ethylhexyl)phthalate
Di -n-Octylphthalate
Benzo (b) f luoranthene
Benzo (k) f luoranthene
Benzo (a) pyrene
Indeno ( 1 , 2 , 3 - cd ) pyrene
Dibenzo (a, h) anthracene
Benzo (g,h, i)perylene
Deuterated Monitoring Compounds
Phenol -d5
bis- ( 2 -Chloroethyl) ether-da
2-Chlorophenol-d4
4-Methylphenol-d8
Nitrobenzene-ds
2 -Nitrophenol -d4
2,4 -Dichlorophenol -d3
4-Chloroaniline-cl,
Dimethylphthalate-d6
Acenaphthylene - dB
4 -Nitrophenol -d4
Fluorene-d10
4, 6-Dinitro-2-methylphenol-d2
Anthracene-d10
Pyrene -d10
Benzo ( a ) pyrene - d12
Minimum RRF
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0

0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
.700
.010
.600
.600
.010
.010
.800
.700
.010
.010
.700
.700
.700
.500
.400
.500

.010
.010
.010
.010
.010
.010
.010
.010
.010
.010
.010
.010
.010
.010
.010
.010
Maximum %RSD
20.5
none
20.5
20.5
none
none
20.5
20.5
none
none
20.5
20.5
20.5
20.5
20.5
20.5

none
npne
none
none
none
none
none
none
none
none
none
none
none
none
none
none
Maximum %Diff
±25.0
none
±25.0
±25.0
none
none
±25.0
±25.0
none
none
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0

none
none
none
none
none
none
none
none
none
none
none
none
none
none
none
none
                              D-49/SV
OLC03.2

-------
Exhibit D Semivolatiles - Section 17
Tables/Diagrams/Flowcharts  (Con't)
                                        Table  D-5
                      Deuterated Monitoring Compound Recovery Limits
              Compound
              Phenol-ds
              bis-(2-Chloroethyl)ether-da
              2-Chlorophenol-dj
              4-Methylphenol-dB
              Nitrobenzene-ds
              2-Nitrophenol-d,
              2,4-Dichlorophenol-d3
              4-Chloroani1ine-d<
              Dimethylphthalate-d6
              Acenaphthylene-da
              4-Nitrophenol-d4
              Fluorene-d10
              4,6-Dinitro-2-methylphenol-d2
              Anthracene -d10
              Pyrene-d10
              Benzo (a) pyrene-d12	
                    % Recovery
                       10-110
                        41-94
                       33-liO
                        38-95
                       35-114
                       40-106
                        42-98
                        8-70
                       62-102
                        49-98
                        9-181
                        50-97
                       53-153
                       55-116
                       47-114
                       54-120
 OLC03.2
D-50/SV

-------
                                                    Exhibit D Semivolatiles -- Section 17
                                                       Tables/Diagrams/Flowcharts (Con't)
Compound
                         Table D-6
Matrix Spike Recovery and Relative Percent Difference Limits

                                         % Recovery
                                                                             RPD
Phenol
2-Chlorophenol
N-Nitroso-di-n-propylamine
4-Chloro-3-methylphenol
Acenaphthene
4-Nitrophenol
2,4-Dinitrotoluene
Pentachlorophenol
Pyrene
                                           12-110
                                           27-123
                                           41-116
                                           23-97
                                           46-118
                                           10-80
                                           24-96
                                           9-103
                                           26-127
42
40
38
42
31
50
38
50
31
                                          D-51/SV
                                                                   OLC03.2

-------
                       EXHIBIT D

METHOD FOR THE ANALYSIS OF LOW CONCENTRATION WATER FOR
                PESTICIDES AND AROCLORS
                        D-l/PEST                                  OLC03.2

-------
                           THIS  PAGE  INTENTIONALLY  LEFT BLANK
OLC03.2                                  D-2/PEST

-------
                 Exhibit D -- Analytical Methods for Pesticides/Aroclors

                                    Table of Contents

                                                                                     Page


1.0    SCOPE AND APPLICATION. 	 5

2.0    SUMMARY OF METHOD  .   . .  . :	6

3.0    DEFINITIONS	6

4.0    INTERFERENCES  	 7
       4.1    Method Interferences	7
       4.2    Matrix Interferences  	7

5.0    SAFETY	7

6.0    EQUIPMENT AND SUPPLIES 	 8
       6.1    Glassware	 8
       6.2    Florisil Cleanup Equipment	 9
       6.3    pH Paper	9
       6.4    pH Meter   	9
       6.5    Boiling Chips  	 9
       6.6    Water Bath	9
       6.7    Balances   	9
       6.8    Nitrogen Evaporation Device	9
       6.9    Mechanical Shaker or Mixer   	9
       6.10   Gas Chromatograph/Electron Capture Detector (GC/ECD) System  	   10

7.0    REAGENTS AND STANDARDS	   13
       7.1    Reagents   	   13
       7.2    Standards	14

8.0    SAMPLE COLLECTION, PRESERVATION, AND STORAGE  	   18
       8.1    Sample Collection and Preservation   	   18
       8.2    Procedure for  Sample Storage   	   18
       8.3    Procedure for  Sample Extract Storage   	   18
       8.4    Records for Sample and Sample Extract Storage  	   18
       8.5    Contract Required Holding Times  	   18

9.0    CALIBRATION AND STANDARDIZATION   	   19
       9.1    Gas Chromatograph  (GC) Operating Conditions  	   19
       9.2    Initial Calibration  	   19
       9.3    Calibration Verification   	   26

10.0   PROCEDURE   	   31
       10.1   Sample Preparation   	   31
       10.2   GC Analysis	   39

11.0   DATA ANALYSIS AND CALCULATIONS	   43
       11.1   Qualitative Identification of Target Compounds   	   43
       11.2   Calculations   	   43
       11.3   Technical Acceptance Criteria for Sample Analyses  	   46
       11.4   Corrective Action for Sample Analysis  ....'.	'	47

12.0   QUALITY CONTROL   	   49
       12.1   Blank Analyses	49
       12.2   Laboratory Control Sample  (LCS)  	   53
       12.3   Matrix Spike/Matrix Spike Duplicate  (MS/MSD)   	.'   54
       12.4   Method Detection Limit  (MDL) Determination   	   57
                                         D-3/PEST                                  OLC03.2

-------
                 Exhibit D -- Analytical Methods for Pesticides/Aroclors





                              Table of Contents  (Continued)





                                                                                     Page








13 . 0   METHOD PERFORMANCE	  58




14 . 0   POLLUTION PREVENTION	'	  58





15.0   WASTE MANAGEMENT  .	  .  58





16.0   REFERENCES	58




17.0   TABLES/DIAGRAMS/FLOWCHARTS 	  59
OLC03.2                          •        D-4/PEST

-------
                                               Exhibit D Pesticides/Aroclors -- Section 1
                                                                    Scope and Application
1. 0    SCOPE AND 'APPLICATION
1.1    The analytical method that follows is designed to analyze water in order
       to determine the presence and concentrations of the chlorinated
       pesticides and Aroclors found in the Target Compound List (Exhibit C -
       Pesticides).   The majority of the samples are expected to be from
       drinking water and well/ground water type sources around Superfund
       sites.   The method can be used for determining analyte concentrations as
       low as  ten parts per trillion.  The method is based on EPA Method 608.
       The method includes sample extraction, extract cleanup techniques, and
       the Gas Chromatograph/Electron Capture Detector (GC/ECD) analytical
       method  for chlorinated pesticides and aroclors.

1.2    Resolution difficulties have been associated with the following pairs of
       compounds using this method:

       •    On a DB-608  or equivalent column,  DDE and Dieldrin,-  Methoxychlor arid
            Endrin ketone;  and Endosulfan I and gamma-Chiordane.

       •    On a DB-1701 or equivalent column,  Endosulfan  I and gamma-Chlordane;
            and Methoxychlor and Endosulfan sulfate.

1.3    There are two isomers of heptachlor epoxide, the endo epoxy isomer
       (isomer A) and the exo epoxy isomer  (isomer B).  The two isomers are
       separable using current GC capillary columns.  Only the exo epoxy isomer
       (isomer B) is of environmental significance.  This is the isomer that
       must be used as an analytical standard, identified and quantitated in
       .sample analysis, and reported on appropriate forms as heptachlor
       epoxide.
                                         D-5/PEST                                  OLC03.2

-------
Exhibit D Pesticides/Arociors -- Sections 2 & 3
Summary of Method

2.0    SUMMARY OF METHOD

2.1    A one liter aliquot of sample is spiked with the surrogate solution and
       extracted with methylene chloride by using a continuous liquid-liquid
       extractor or separatory funnel.   The methylene chloride extract is dried
       and concentrated, exchanged to hexane,  cleaned up to remove
       interferences, and adjusted to a final volume of 2.0 milliliters (mL).

2.2    The hexane extract is injected onto two wide-bore capillary columns in a
       Gas Chromatograph (GO.   The GC is temperature programmed to separate
       the pesticides and Aroclors which are then detected with an Electron
       Capture Detector  (BCD).   Calibration and run sequence specifications of
       the GC/ECD method apply independently to each GC column.

2.3    A single component pesticide is identified if a peak is detected within
       its appropriate Retention Time (RT) window on each of two GC columns.
       Quantitative analysis of pesticides/Aroclors must be accomplished by the
       external standard method.  Single component analytes and the surrogates
       must be analyzed at three concentration levels during the initial
       calibration.

2.4    Toxaphene and Aroclors are identified primarily by pattern recognition,
       but RTs of three to five major peaks must also be taken into
       consideration.  Single-point calibrations for multicomponent analytes
       are sufficient for quantitation by this method.  Standards for
       identified Aroclors and Toxaphene must be run within 72 hours of the
       sample analysis in which they were observed.  These standards are used
       to verify identification only; quantitation is based on the standards
       analyzed during initial calibration.

3.0    DEFINITIONS

       See Exhibit G for a complete list of definitions.
OLC03.2                                  D-6/PEST

-------
                                          Exhibit D Pesticides/Aroclors -- Sections 4 & 5
                                                                            Interferences

4.0    INTERFERENCES

4 .1    Method Interferences

       Method interferences may be caused by contaminants in solvents,
       reagents,  glassware, and sample processing hardware.   These contaminants
       lead to discrete artifacts or to elevated baselines in gas
       chromatograms.   These materials must be routinely demonstrated to be
       free from interferences under the sample preparation and analysis
       conditions by analyzing instrument blanks and method blanks.
       Interferences caused by phthalate esters can pose a major problem in
       pesticide analysis.  Because common flexible plastics contain varying
       amounts of phthalates which are easily extracted during laboratory
       operations, cross-contamination of glassware frequently occurs when
       plastics are handled.  Interferences from phthalates can best be
       minimized by avoiding the use of such plastics in the laboratory.

4.2    Matrix Interferences

       Matrix interferences may be caused by contaminants that are co-extracted
       from the sample.  The extent of matrix interferences will vary
       considerably from source to source, depending upon the nature of the
       site being sampled.  The cleanup procedures in this method must be used
       to remove such interferences in order to achieve the Contract Required
       Quantitation Limits  (CRQL).

5.0    SAFETY

5.1    The toxicity or carcinogenicity of each reagent used in this method has
       not been precisely determined; however, each chemical should be treated
       as a potential health hazard.  Exposure to these reagents should be
       reduced to the lowest possible level.  The laboratory is responsible for
       maintaining a current awareness file of OSHA regulations regarding the
       safe handling of the chemicals specified in this method.  A reference
       file of data handling sheets should be made available to all personnel
       involved in these analyses. -Specifically, concentrated sulfuric acid
       and the 10 N sodium hydroxide solution are moderately toxic and
       extremely  irritating to skin and mucous membranes.  Use these reagents
       in a fume hood whenever possible and if eye or skin contact occurs,
       flush with large volumes of water.  Always wear safety glasses or a
       shield for eye protection, protective clothing, and observe proper
       mixing when working with these reagents.

5.2    The following analytes covered by this method have been tentatively
       classified as known or suspected human or mammalian carcinogens: 4,4'-
       DDT, 4,4'-DDD, the BHCs, and the Aroclors.  Primary standards of these
       toxic compounds should be prepared in a hood.  A NIOSH/MESA approved
       toxic gas  respirator should be worn when the analyst handles high
       concentrations of  these toxic compounds.
                                         D-7/PEST                                  OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 6
Equipment and Supplies

6.0    EQUIPMENT AND SUPPLIES

       Brand names, suppliers, and part numbers are for illustrative purposes
       only.  No endorsement is implied.  Equivalent performance may.be
       achieved using equipment and supplies other than those specified here,
       but demonstration of equivalent performance meeting the requirements of
       the Statement of Work  (SOW) is the responsibility of the Contractor.
       The Contractor must document in its Sample Delivery Group (SDG)
       Narrative when it uses equipment and supplies other than those specified
       here.

6.1    Glassware

6.1.1     Continuous  Liquid-Liquid Extractors

          Continuous  Liquid-Liquid Extractors  equipped with  PTFE or glass
          connecting  joints  and stopcocks  requiring no lubrication (Hershberg-
          Wolf  extractor,  Ace Glass  Company, Vineland, NJ  P/N 6841-10 or
          equivalent)  or  Hydrophobic Membrane-based Extractor (Accelerated One
          Step™ Extractor, Corning series  3195 or equivalent).

6.1.2     Separatory  Funnels -  2  liter with PTFE  stopcock. '

6.1.3     Graduated Cylinder -  1  liter capacity.

6.1.4     Drying column,  chromatographic column approximately. 400  millimeters
           (mm)  long x 19  mm  ID,  with coarse frit.   (Substitution of a small pad
          of  disposable Pyrex glass  wool for the  frit  will help prevent cross-
          contamination of sample extracts.)

6.1.5     Kuderna-Danish  Apparatus

6.1.5.1       Concentrator Tubes - 10 milliliters  (mL), graduated  (Kontes,
              K-570050-1025,  or equivalent).

6.1.5.2       Evaporative Flasks - 500 mL  (Kontes K-570001-0500,  or equivalent).
              Attach to concentrator tube with springs.

6.1.5.3       Snyder Columns - three-ball macro (Kontes K-50300-0121, or
              equivalent).

6.1.5.4    •   Snyder Columns, micro two or three ball with a 19 mm ground glass
              joint.

6.1.6     Pipet,  Volumetric  1.00  mL  or 2.00 mL.

6.1.7     Microsyringe, 1.0  microliter (uL)  and larger,  0.006 (0.15 mm) inch ID
          needle.

6.1.8     Syringe,  1.00 mL or 2.00 mL (optional).

6.1.9     Volumetric  flask,  10.00 mL,  and  1 or 2  mL.

6.1.10    Vials and caps,  20 and 10  mL, with screw cap and PTFE or aluminum
          foil  liner,  2 or 1 mL for  Gas Chromatograph  (GC) auto sampler.

6.1.11    Bottle or test  tube,  20 mL with  PTFE-lined screw cap for sulfur
          removal.

6.1.12    Centrifuge  tubes,  calibrated, 12 mL,  for sulfur removal.

6.1.13    Micropipet,  200 uL, with disposable  tips.


OLC03.2                                  D-8/PEST

-------
                                               Exhibit D Pesticides/Aroclors — Section 6
                                                           Equipment and Supplies (Con't)
6.2    Florisil Cleanup Equipment
6.2.1     Florisil bonded silica.  1 g cartridges with stainless steel or PTFE
          frits. Catalog No. 694-313  (Analytichem, 24201 Frampton Ave., Harbor
          City, CA, or equivalent).

6.2.2     Vacuum system for eluting multiple cleanup cartridges.  Vac Elute
          Manifold, Analytichem  International, J.T. Baker, or Supelco  (or
          equivalent).

6.2.3     Vacuum trap made from  a  500 mL sidearm flask fitted with a one-hole
          stopper and glass tubing.  -

6.2.4     Vacuum pressure gauge.

6.2.5     Rack  for holding 10  mL volumetric flasks in the manifold.

          NOTE: Other types of equivalent  systems, such as an automated system
          using syringe pressure are  considered acceptable for  elution of
          Florisil cartridges, as  long as  all quality control  (QC) and sample
          technical acceptance criteria are met.

6.3    pH Paper, Wide Range

6.4    pH Meter — - With a combination glass electrode.  Calibrate according to
       manufacturer's instructions.  pH meter must be calibrated prior to each
       use.

6.5    Boiling Chips

6.5.1     Silicon  carbide boiling  chips, approximately 10 to 40 mesh.  Heat  the
          chips to 400°C  for  30  minutes or solvent rinse before use.

6.5.2     PTFE  boiling  chips  (optional).   Solvent  rinse the  chips before use.

6.6    Water Bath, heated, with  concentric ring cover, capable  of temperature
       control.

       NOTE: To prevent the release of solvent fumes into the laboratory, the
       water bath must be used in  a hood.

6.7    Balances

       Analytical balances, capable of weighing accurately to ±0.0001 g.  The
       balances must be calibrated with class S weights or known reference
       weights  once per each  12-hour  work  shift.  The balances  must be
       calibrated with class  S weights at  a minimum of once  per month.  The
       balances must also be -annually checked by a certified technician.

6.8    Nitrogen Evaporation Device

       Nitrogen evaporation device equipped with a heated bath  that can be
       maintained  at  35 to 40°C  (N-Evap by Organomation Associates, Inc., South
       Berlin,  MA, or equivalent).  To prevent the release of solvent fumes
       into the laboratory, the  nitrogen evaporation device  must be used in a
       hood.

6.9    Mechanical  Shaker or Mixer, for Sulfur Removal
                                         D-9/PEST                                  OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section S
Equipment and Supplies  (Con't)

6.10   Gas Chromatograph/Electron Capture Detector (GC/ECD) System

6.10.1    Gas Chromatograph  (GC)

6.10.1.1      The GC must adequately regulate temperature in order to give a
              reproducible temperature program and have a flow controller that
              maintains a constant column flow rate throughout temperature
              program operations.   The system must have all required accessories
              including syringes,  analytical columns,  and gases.

6.10.1.2      GCs that are available from some manufacturers may have difficulty
              in meeting certain method QC requirements because of endrin and
              DDT breakdown in the injector.  This problem can be minimized by
              operating the injector at 200-205°C,  using a Pvrex (not quartz)
              methyl silicone deactivated injector liner, and deactivating the
              metal parts in the injector with dichlorodimethyl silane.  In some
              cases, using a 0.25 inch packed column injector converted for use
              with 0.53 mm capillary columns works better than a Grob-type
              injector.  If a Grob-type injector is used, a 4 mm liner may be
              required to meet breakdown criteria.

6.10.1.3      Gas Chromatograph Columns .-- Two wide-bore (0.53 mm ID) fused
              silica GC columns are required.  A separate detector is required
              for each column.  The specified analytical columns are a 30 m x
              0.53 mm ID, 1.0 urn film thickness DB-1701  (J&W Scientific); SPB
              1701  (Supelco); AT 1701  (Alltech); RTX-1701  (Restek); CP-Sil 19CB
              (Chrompack); 007-1701 (Quadrex); BP-10 (SGE); or'equivalent, and a
              30 m x 0.53 mm ID, 0.5 to 1.0 micrometer (• m)  film thickness DB-
              608 (J&W Scientific);  HP-608  (Hewlett Packard); SPB-608  (Supelco);
              007-608  (Quadrex); BP-608 (SGE); CP-Sil 8CB  (Chrompack); or
              equivalent.

              NOTE: The column length stated above is the minimum requirement.
              Longer columns that meet resolution and calibration requirements
              may be used.  A description of the GC columns used for analysis
              shall be provided in the SDG Narrative.

6.10.1.3.1        A  capillary column is considered equivalent  if:

                  •    The column does not introduce contaminants which, interfere
                      with identification and quantitation of the compounds
                      listed in Exhibit C (Pesticides).

                  •    The analytical results generated using the column meet the
                      initial calibration and calibration verification technical
                      acceptance criteria listed in the SOW and the Contract
                      Required Quantitation Limits (CRQLs)  listed in Exhibit C
                       (Pesticides).

                  •'   The column can accept at least 16 times the low point
                      standard for individual standard mixtures A and B for each
                      compound listed in Exhibit C (Pesticides) without becoming
                      overloaded.

                  •    The column pair chosen must have dissimilar phases/chemical
                      properties in order to separate the compounds of interest
                      in different Retention Time  (RT)  order.

6.10.1.3.2        Although the instructions  included  in the SOW  are  for wide bore
                  capillary columns, narrower bore capillary columns may  be
                  evaluated for  use.   Follow manufacturer's instructions  for use
OLC03.2                                  D-10/PEST

-------
                                               Exhibit D Pesticides/Aroclors — Section 6
                                                           Equipment and Supplies (Con't)

                 of its product.  Document in SDG Narrative if other columns are
                 used'by specifying the column used (Exhibit B Section 2.5.1) .

6.10.1.3.3       As applicable, follow the manufacturer's instructions for use
                 of its product.

6.10.1.3.4       The Contractor must maintain documentation that the alternate
                 column met the criteria in Sections 9.2.5 and 9.3.5.  The
                 minimum documentation is as follows:

6.10.1.3.4.1        Manufacturer provided  information concerning the performance
                    characteristics  of  the column;

6.10.1.3.4.2        GC chromatograms and data  system reports  generated on the
                    GC/ECD and used  for CLP analyses:

                     •  From instrument blanks which demonstrate that there are
                        no contaminants which interfere with the pesticide
                        analysis  when using the alternate column;

                     •  For initial calibration standards analyzed using the
                        alternate column; and

                     •  For calibration verification standards analyzed using
                        the alternate column.

6.10.1.3.5       Based on  the  Contractor generated data described in Section
                 6.10.1.3.4.2,  the Contractor must complete a written comparison
                 and review, signed  by  the  Laboratory  Manager certifying  that:

                 •    The alternate column performance is comparable to the
                      required column performance in its ability to produce
                      initial calibrations and calibration verifications which
                      meet the technical acceptance criteria in Sections 9.2.5
                      and 9.3.5;

                 •    The low point initial calibration standard analyses have
                      adequate sensitivity to meet the pesticide CRQLs;

                 •    The high point initial calibration standard analyses were
                      not overloaded; and

                 •    The alternate column does not introduce contaminants which
                      interfere with identification and quantitation of compounds
                      listed in Exhibit C  (Pesticides).


6.10.1.3.6       The  documentation must be made  available  to  USEPA  during on-
                 site laboratory evaluations or  sent  to USEPA upon  request of
                 the  Contract  Laboratory Program Project Officer (CLP PO).

6.10.1.3.7       PACKED COLUMNS CANNOT  BE  USED.

6.10.1.4      Columns are mounted in a 0.25-inch injector ports by using glass
              adapters  available  from a variety of commercial sources  (J&W
              Scientific, Supelco, Inc., Hewlett-Packard, Varian, Inc.,  Perkin
              Elmer, or equivalent).  The two columns may be mounted  into a
              single injection port with a tee adapter (Supelco, Inc.,
              Bellefonte, PA,  Catalog No. 2-3660, or equivalent).  Use of this
              adapter  allows simultaneous injection onto both columns.  The
              laboratory should  follow manufacturer's  recommendations  for
              mounting  0.53 mm capillary columns in injector ports.


                                         D-11/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors — Section 6
Equipment and Supplies  (Con't)

6.10.1.5      The carrier gas for routine applications is helium.  Laboratories
              may choose to use hydrogen as a carrier gas, but they must clearly
              identify its use in the SDG Narrative and on all divider pages
              preceding raw chromatographic data in submissions to USEPA.
              Laboratories that choose to use hydrogen are advised to exercise
              caution in its use.  Use of a hydrogen leak detector is highly
              recommended when hydrogen is used as the carrier gas.  All GC
              carrier gas lines must be constructed from stainless steel or
              copper tubing.  Non-polytetrafluoroethylene (PTFE)  thread sealants
              or flow controllers with rubber components are not to be used.

6.10.2    Electron  Capture  Detector (BCD)

          The  linearity of  the  response  of  the  BCD may be greatly dependent on
          the  flow  rate of  the  make-up gas.   The  make-up gas  must be P-5,  P-10
           (argon/methane) or nitrogen according to the instrument
          specification.  Care  must be taken to maintain stable  and appropriate
          flow of make-up gas to the detector.  The  GC/ECD  system must  be  in a
          room in which the atmosphere has  been demonstrated  to  be free of all
          contaminants  which may interfere  with the  analysis.  The instrument
          must be vented to outside the  facility  or  to a trapping system which
          prevents  the  release  of contaminants  into  the instrument room.

6.10.3    Data System

          A data system must be interfaced  to the GC/ECD.  The data system must
          allow the continuous  acquisition  of data throughout the duration of
          the  chromatographic program and must  permit,  at the minimum,  the
          output of time vs. intensity (peak height  or peak area)  data.  Also,
          the  data  system must  be able to rescale chromatographic data  in  order
          to report chromatograms meeting the requirements  listed within this
          method.

6.10.4    Magnetic  Tape Storage Device

          Magnetic  tape storage devices  must be capable of  recording data  and
          suitable  for  long-term,  off-line  storage of  GC/ECD  data.
OLC03.2                                 D-12/PEST

-------
                                               Exhibit D Pesticides/Aroclors -- Section 7
                                                                   Reagents and Standards

7.0    REAGENTS AND STANDARDS

7.1    Reagents

       Reagents shall be dated with the receipt date and used on a first-in,
       first-out basis.  The purity of the reagents shall be verified before
       use.

7.1.1     Reagent water  -- Defined as water in which no  target analyte is
          observed  at  the Contract Required Quantitation Limits  (CRQL) for that
          compound.

7.1.1.1       Reagent water may be generated by passing tap water through a
              carbon filter bed containing about 453 grams (g)  (1 pound)  of
              activated carbon (Calgon Corp.,  Filtrasorb-300,  or equivalent).

7.1.1.2       Reagent water may be generated using a water purification system
              (Millipore Super-Q or equivalent).

7.1.2     Sodium sulfate, granular-anhydrous  reagent grade, heated  at 400°C for
          4  hours,  or at 120°C for 16  hours,  cooled in a desiccator, and stored
          in a glass bottle.   Each lot  must be  extracted with hexane and
          analyzed by Gas Chromatograph/Electrdn Capture Detector (GC/ECD) to
          demonstrate that  it is free  of  interference  before use.   J. T.  Baker
          anhydrous granular, Catalog  No.  3375,  or equivalent.

          CAUTION:  An open  container of sodium sulfate may become contaminated
          during storage in the laboratory.

7.1.3     Methylene chloride, hexane,  acetone,  toluene,  iso-octane,  and
          methanol (optional), pesticide  quality,  or equivalent.  It is
          recommended that  each lot  of solvent  be analyzed to demonstrate that
          it is free of interference before  use.  Methylene chloride must be
          certified as acid free or  must  be  tested to demonstrate that it is
          free of hydrochloric acid.  Acidic methylene chloride must be  passed
          through basic alumina and  then demonstrated to be free of
          hydrochloric acid.

7.1.4     Mercury,  triple distilled, for sulfur clean-up.

7.1.5     Copper powder (optional),  fine,  granular (Mallinckrodt 4649  or
          equivalent).  Copper may be used instead of mercury for sulfur clean-
          up.  Remove oxides by treating with dilute nitric acid, rinse  with
          distilled water to remove  all traces of acid,  rinse with  acetone, and
          dry under a stream of nitrogen.

7.1.6     Sodium hydroxide solution  (10 N).   Carefully dissolve 40  g of  NaOH  in
          reagent water and dilute the solution to 100 milliliters  (mL).

7.1.7     Concentrated sulfuric acid,   (Sp. Gr.  1.84)-36N.

7.1.8     Nitric acid, dilute, for sulfur removal with copper.

7.1.9     Ten percent acetone in hexane (v/v).   Prepare by adding 10.0 mL of
          acetone to  90.0 mL of hexane.

          NOTE: Prepare this mixture accurately or the results  from the
          Florisil cartridge cleanup will be adversely affected. Water  in the
          acetone will also adversely affect Florisil performance.
                                         D-13/PEST                                 OLC03.2

-------
Exhibit  D  Pesticides/Aroclors --  Section 7
Reagents and Standards (Con't)

7.2    Standards

       The  Contractor  must provide  all  standards  to be used with  this  contract.
       These standards may be used  only, after  they have been  certified
       according  to  the procedure in Exhibit E.   The Contractor must be  able  to
       verify that the standards  are certified.   Manufacturer's certificates  of
       analysis must be retained  by the Contractor and presented  upon  request.

7.2.1      Stock Standard Solutions

7.2.1.1        Stock  standard  solutions  may be  purchased as certified solutions
               or  prepared from pure standard materials.

7.2.1.2        Prepare  stock standard solutions by accurately  weighing  about
               0.0100 g of pure material.  Dissolve the material in toluene,
               dilute to volume in a 10  mL volumetric flask with toluene  or
               acetone.  Larger volumes  may be  used at the convenience  of the
               analyst.

7.2.1.3        When compound purity  is assayed  to  be 97 percent or greater, the
               weight may be used  without  correction to calculate  the
               concentration of the  stock solution.  If the compound purity is
               assayed  to be less  than -97  percent, the weight  must be corrected
               when calculating the  concentration  of the stock solution.

7.2.1.4        Fresh  stock standards must  be prepared once every six months, or
               sooner,  if standards  have degraded  or concentrated.  Stock
               standards must  be checked for signs of degradation  or
               concentration just  prior  to preparing working standards  from them.

7.2.2      Secondary Dilution Standards

7.2.2.1        Using  stock standards, prepare secondary dilution standards in
               acetone  that contain  the  compounds  of interest  either singly or
               mixed  together.

7.2.2.2        Fresh  secondary dilution  standards  must be prepared once every  six
               months,  or sooner,  if standards  have degraded or concentrated.
               Secondary dilution  standards must be checked for signs of
               degradation or  concentration just prior to preparing working
               standards from  them.

7.2.3      Working Standards

.7.2.3.1        Surrogate Standard  Spiking Solution

               The surrogates, tetrachloro-m-xylene and decachlorobiphenyl, are
               added  to all standards, samples  (including Laboratory Control
               Samples), matrix spike, matrix spike duplicates and if required,
               Performance Evaluation  (PE) samples, and required blanks
               (method/sulfur  clean-up/instrument).  Prepare a surrogate  spiking
              .solution of 0.20 micrograms per  milliliter  (ug/mL)  of each of the
               two compounds in acetone.   The solution should  be checked
               frequently for  stability.   The solution must be replaced every  6
               months,  or sooner if  the  solution has degraded  or concentrated.

 7.2.3.2        Matrix Spiking  Solution

               Prepare  a matrix spiking  solution in acetone or methanol that
               contains the following pesticides at the concentrations  specified.
               The solution must be  replaced every six months, or  sooner  if the
               solution has degraded or  concentrated.


OLC03.2  '                                D-14/PEST

-------
7.2.3.3
                                               Exhibit D Pesticides/Aroclors -- Section 7
                                                           Reagents and Standards (Con't)
 Pesticide
 gamma- BHC  (Lindane)
 4,4'-DDT
 Endrin
 Heptachlor
 Aldrin
 Dieldrin
Resolution Check Mixture
                                                     • Concentration (uq/mL)
                                                               0.1
                                                               0.2
                                                               0.2
                                                               0.1
                                                               0.1
                                                               0.2
              The Resolution Check Mixture is composed of the pesticides and
              surrogates at the concentrations listed below in hexane or iso-
              octane.   The mixture must be prepared every six months, or sooner,
              if the solution has degraded or concentrated.
               Compounds
               gamma- Chlordane
               Endosulfan I
               4,4'-DDE
               Dieldrin
               Endosulfan sulfate
               Endrin ketone
               Methoxychlor
               Tetrachloro-m-xylene
               Decachlorobiphenyl
                                        Concentration  (na/mL)
                                                 10.0
                                                 10.0
                                                 20.0
                                                 20.0
                                                 20.0
                                                 20.0
                                                 100.0
                                                 20.0
                                                 20.0
7.2.3.4       Performance Evaluation Mixture (PEM)
              The PEM is prepared in hexane or iso-octane, as listed below.
              PEM must be prepared weekly, or sooner if the solution has
              degraded or concentrated.
                                                               The
               Compounds
               gamma-BHC
               alpha-BHC
               4,4'-DDT
               beta-BHC
               Endrin
               Methoxychlor
               Tetrachloro-m-xylene
               Decachlorobiphenyl
                                         Concentration  (nq/mL)
                                                 10.0
                                                 10.0
                                                 100.0
                                                 10.0
                                                 50.0
                                                 250.0
                                                 20.0
                                                 20.0
                                         D-15/PEST
                                                                                   OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 7
Reagents and Standards  (Con't)
7.2.3.5       Individual Standard Mixtures A and B
              The Individual Standard Mixture solutions must be prepared in
              either hexane or iso-octane.  The concentrations of the pesticides
              in the low point standard mixtures are given below.  The midpoint
              concentration must be 4 times the low point concentration for each
              analyte,  including the surrogates.  The high concentration must be
              at least 16 times the low point concentration for each analyte,
              including the surrogates, but a higher concentration may be chosen
              by the Contractor.  The high point concentration defines the upper
              end of the concentration range for which the calibration is valid.
              The solution must be prepared every six months, or sooner, if the
              solution has degraded or concentrated.
 Individual Standard
 Mix A
  Low Point
Concentration
   (ng/mL)
Individual Standard
Mix B
Low Point
Concentration
(ng/mL)	
 alpha-BHC
 Heptachlor
 gamma-BHC
 Endosulfan  I

 Dieldrin
 Endrin
 4,4'-DDD
 4,4'-DDT
 . Methoxychlor
 Tetrachloro-m-xylene
 Decachlorobiphenyl
     5.0        beta-BHC
     5.0        delta-BHC
     5.0        Aldrin
     5.0        Heptachlor-epoxide
                (exo-epoxy isomer)
     10.0        alpha-Chiordane
     10.0        gamma-Chiordane
     10.0        4,4'-DDE
     10.0        Endosulfan sulfate
     50.0        Endrin aldehyde
     5 .0        Endrin ketone
     10.0        Endosulfan II
                Tetrachloro-m-xylene
                Decachloro-biphenyl
                                   5.0
                                   5.0
                                   5.0
                                   5.0

                                   5.0
                                   5.0
                                   10.0
                                   10.0
                                   10:0
                                   10.0
                                   10.0
                                   5.0
                                   10.0
  NOTE:  Only the  exo-epoxy isomer (isomer B)  of  heptachlor  expoxide  is  used as
  an analytical standard.
7.2.3.6      Multicomponent Standards
              Toxaphene and Aroclor standards must be prepared individually
              except for Aroclor 1260 and Aroclor 1016 which may be combined in
              one standard mixture.  The calibration standards for the Aroclors
              must be prepared at concentrations of 100 nanograms (ng)/mL,
              except for Aroclor 1221 which must be prepared at 200 ng/mL.
              Toxaphene must be prepared at 500 ng/mL.  All multicomponent
              standards must contain the surrogates  at 20.0 ng/mL.   The Aroclor
              and toxaphene solutions must be prepared in hexane or iso-octane.
              Each solution must be prepared every 6 months, or sooner, if the
              solution has degraded or concentrated.
7.2.3.7      Florisil Cartridge Check Solution
              Prepare a 0.10 ug/mL solution of 2,4,5-trichlorophenol in acetone.
              The solution must be prepared every six months, or sooner, if the
              solution has degraded or concentrated.
OLC03.2
            D-16/PEST

-------
                                               Exhibit D Pesticides/Aroclors -- Section 7
                                                           Reagents and Standards (Con't)

7.2.3.8       Laboratory Control Sample (LCS)  Spiking Solution

              Prepare a LCS spiking solution that contains  each of  the analytes
              at the concentrations- listed below in methanol or acetone.   The
              LCS solution must be prepared every six months,  or sooner,  if the
              solution has degraded or concentrated.

                  Compounds                                    Concentration (uq/mL)
                  gamma-BHC                                            0.10
                  Heptachlor epoxide                   .                0.10
                  Dieldrin                                             0.20
                  4,4'-DDE                                             0.20
                  Endrin                                               0.20
                  Endosulfan sulfate                                   0.20
                  gamma-Chlordane                                      0.10

7.2.4     Ampulated Standard Extracts

          Standard solutions purchased from a chemical supply house as
          ampulated extracts in glass  vials may be retained and used  until the
          expiration date provided by the manufacturer.  If no manufacturer's
          expiration date is provided,  the standard solutions as ampulated
          extracts may be retained and used for 2 years  from the preparation
          date.   Standard solutions prepared by the Contractor which  are
          immediately ampulated in glass  vials may be retained for 2  years from
          preparation date.   Upon breaking the glass seal,  the expiration times
          listed in Sections 7.2.1 to 7.2.3 will apply.  The Contractor  is
          responsible for assuring that the integrity of the standards has not
          degraded (Section 7.2.5.5).

7.2.5     Storage of Standards

7.2.5.1       Store the stock and  secondary.standard solutions at 4°C (±2°C)  in
              PTFE-lined,  screw-cap, amber bottles/vials.

7.2.5.2       Store the working  standard  solutions at 4°C  (±2°C) in PTFE-lined
              screw-cap,  amber bottles/vials.  The working  standards must be
              checked  frequently for signs of degradation or evaporation.

7.2.5.3       Protect  all standards from  light.

7.2.5.4.      Samples,  sample extracts, and standards must  be  stored separately.

7.2.5.5       The  Contractor is  responsible for maintaining the  integrity of
              standard solutions and verifying prior to use.   Storage of
              standard solutions in the freezer may cause some compounds to
              precipitate.   This means at the minimum, the  standards must be .
              brought  to  room temperature prior to use,  checked  for losses, and
              checked  that all  components have remained in  solution.  Additional
               steps may be necessary to ensure all components  are  in  solution.

7.2.6      Temperature Records for Storage of Standards

7.2.6.1       The  temperature of all standard storage refrigerators/freezers
               shall be recorded daily.

7.2.6.2       Temperature excursions shall  be noted  and appropriate corrective
               actions  shall  be  taken to correct problems, when required.
                                         D-17/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Sections 7 6 S
Sample Collection, Preservation, and Storage

7.2.6.3       Corrective action SOPs shall be posted on the refrigerators.

8.0    SAMPLE COLLECTION, PRESERVATION, AND STORAGE

8.1    Sample Collection and Preservation

8.1.1     Water samples  may be  collected  in  1 L  (or  1  quart)  amber glass
          containers,  fitted with  screw-caps  lined with  PTFE.   If amber
          containers are not available, the  samples  should be protected from
          light.  The  specific  requirements  for  site sample collection are
          outlined by  the Region.

8.1.2     All samples  must  be iced or  refrigerated at  4°C  (±2°C) from the time
          of collection  until extraction.

8.2    Procedure for Sample Storage

8.2.1     The samples  must  be protected from light and refrigerated  at 4°C
           (±2°C)  from  the time  of  receipt until  60 days  after delivery of a
          complete reconciled sample data package to USEPA.   After 60 days the
          samples may  be disposed  of in a manner that  complies with  all
          applicable regulations.

8.2.2     The samples  must  be stored in an atmosphere  demonstrated to be free
          of all  potential  contaminants.
 ..,-;
8". 3    Procedure for Sample Extract Storage

8:3.1     Sample  extracts must  be  protected  from light and stored at 4°C  (±2°C)
          until 365 days after  delivery of a  complete  reconciled data package
 5        to USEPA.

8.3.2     Sample  extracts must  be  stored  in  an atmosphere  demonstrated to be
          free of all  potential contaminants.

8"3.3     Samples, sample extracts, and standards must be  stored separately.

8.4    Records for Sample and Sample Extract Storage

8.4.1     The temperature of all sample and  sample extract storage
          refrigerators  shall be recorded daily.

8.4.2     Temperature  excursions shall be noted  and  appropriate corrective
          actions shall  be  taken to correct problems,  when required.

8.4.3     Corrective action SOPs shall be posted on  the  refrigerators.

8.5    Contract Required Holding Times

8.5.1     Extraction of  water samples  by  separatory  funnel procedures must be
          completed within  5 days  of the  Validated Time  of Sample Receipt
           (VTSR).  Extraction of water samples by continuous liquid-liquid
          extraction must be started within  5 days of  VTSR.

8.5.2     As part of USEPA's QA program,  USEPA may provide Performance
          Evaluation  (PE) Samples  as standard extracts which the Contractor is
          required to  prepare per  instructions provided  by USEPA.  PE samples
          must be prepared  and  analyzed concurrently with  the samples in the
          SDG.  The extraction  holding time  (5 days  after  VTSR)  does not apply
          for PE  Samples received  as standard extracts.
OLC03.2                                 D-18/PEST

-------
                                          Exhibit D Pesticides/Aroclors -- Sections 8 & 9
                                                          Calibration and Standardization

8.5.3     Analysis of sample extracts must be completed within 40 days
          following the start of extraction.

9.0    CALIBRATION AND STANDARDIZATION

9.1    Gas Chromatograph (GC)  Operating Conditions

9.'1.1     The  following are the gas chromatographic analytical conditions.  The
          conditions are recommended unless otherwise noted.


            Carrier Gas:                 Helium (Hydrogen may be used,  Section
                                        6.10.1.5)

            Column Flow:                 5 mL/min

            Make-up Gas:                 Argon/Methane (P-5 or P-10)  or N2

                                         (required)

            Injector  Temperature:        > 200°C (Section 9.1.5)

            Injection Technique:         On-column

            Injection Volume:            1 or 2 pi (Section 9.1.3)

            Injector:                    Grob-type, splitless

            Initial  Temperature:         150°C

            Initial  Hold Time:           0.5 min

            Temperature Ramp:            5°C to 6°C/min

            Final Temperature:           275°C

            Final Hold Time:             After Decachlorobiphenyl has eluted
                                         (approximately 10 minutes)

9.1.2      Optimize  GC conditions for analyte  separation and sensitivity.   Once
           optimized,  the same  GC conditions must  be used for the analysis of
           all standards,  samples (including Laboratory Control Samples,
           requested matrix spike,  and matrix  spike duplicate),  and  required
           blanks (method/sulfur clean-up/instrument).

9.1.3      Manual injections  must be 2.0 microliter (uL).   Auto injectors  may
           use 1.0 uL volumes.   The same injection volume must be used for all
           standards,  samples (including Laboratory Control Samples,  requested
           matrix spike,  and matrix spike duplicate)  and required blanks
            (method/sulfur clean-up/instrument).

9.1.4      The linearity of the Electron Capture Detector (BCD)  may be greatly
           dependent on the flow rate of the make-up gas.   Care must be taken to
           maintain stable and appropriate flow of make-up gas to the detector.

9.1.5      Cold  (ambient temperature)  on-column injectors that allow injection
           directly onto a 0.53 mm ID column may be used as long as  the initial
           calibration and calibration verification technical acceptance
           criteria are met.

9.2     Initial Calibration

9.2.1      Summary of Initial Calibration

           Prior to sample (including Laboratory Control Samples, requested
           matrix spike and matrix spike duplicate)  and required blanks


                                         D-19/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors  --  Section  9
Calibration and Standardization  (Con't)

           (method/sulfur clean-up)  analysis,  each GC/ECD system must be
           initially calibrated at a minimum of three concentrations for single
           component analytes and surrogates in order to determine instrument
           sensitivity and the linearity of  GC response.  Multicomponent target
           analytes  are calibrated at a single point.

9.2.2      Frequency of Initial Calibration

           Each GC/ECD system must be initially calibrated upon award of the
           contract, whenever major instrument maintenance or modification is
           performed (e.g.,  column replacement or repair, cleaning or
           replacement of BCD,  etc.), or if  the calibration verification
           technical acceptance criteria have not been met.

9.2.3      Procedure for Initial Calibration

9.2.3.1       Set up the GC/ECD system as described  in  Section 9.1.

9.2.3.2       Prepare the initial calibration  standards using the procedures,
              the analytes, and the  concentrations according to Section 7.2.

9.2.3.3       All standards, samples  (including Laboratory Control Samples,
              requested matrix spike, and matrix spike  duplicate) and required
              blanks  (method/sulfur  clean-up)  extracts must be allowed to warm
              to ambient temperature before analysis.

9.2.3.4       Analyze the initial calibration  sequence  as given below.

              NOTE:  Steps 16 and  17  are used as part of  the calibration
              verification as well  (Section 9.3).

                            INITIAL CALIBRATION SEQUENCE
   J                            1 •  Resolution Check
                                2.  Performance Evaluation Mixture
   :.••'"'•                     .  3.  Aroclor 1016/1260
                                4.  Aroclor .1221
                                5.  Aroclor 1232
                                6.  Aroclor 1242
                                7.  Aroclor 1248
                                8.  Aroclor 1254
                                9.  Toxaphene
                                10.  'Low Point Standard A
                                11.  Low Point Standard B
                                12.  Midpoint Standard A
                                13.  Midpoint Standard B
                                14.  High Point Standard A
                                15.  High Point Standard B
                                16.  Instrument Blank
                                17.  Performance Evaluation Mixture

9.2.4      Calculations for Initial Calibration

9.2.4.1       During the initial  calibration sequence,  absolute Retention Times
               (RTs)  are determined for all single component pesticides, the
              surrogates, and at  least three major peaks of each multicomponent
              analyte.


OLC03.2                                  D-20/PEST

-------
                                               Exhibit D Pesticides/Aroclors -- Section 9
                                                  Calibration and Standardization  (Con't)

9.2.4.2       For each single component pesticides,  a RT is measured in each of
              the three calibration standards (low point,  midpoint,  high point)
              for Individual Standard Mixture A and  Individual  Standard Mixture
              B.   The RT for the surrogates is measured from each of the three
              analyses of Individual Standard Mixture A during  the initial
              calibration.  The mean RT is calculated for each  single component
              pesticide and surrogate as the average of the three values.
              Calculate a mean absolute RT for each  single component pesticide
              and surrogate using Equation 1.

              EQ. 1

                                                RT.
              Where:


               RT  =       Mean absolute retention time of analyte.

               RTj =        Absolute retention time of analyte.

               n   =       Number of measurements  (3).

9.2.4.3       A RT window is calculated for each single component analyte and
              surrogate and for the major peaks  (3 to 5) of each multicomponent
              analyte by using Table D-l.  Windows are centered around the
              average absolute RT for the analyte established during the initial
              calibration.  Analytes are: identified when peaks are observed in
              the RT window for the compound on both GC columns.

9.2.4.4       The linearity of the instrument is determined by calculating a
              percent relative standard deviation  (%RSD) of the calibration
             . factors from a three-point calibration curve for each of the
              single component pesticide and surrogates.  Either peak area or
              peak height may be used to calculate calibration factors used in
              the %RSD equation.  For example, it is permitted to calculate
              linearity for endrin based on peak area and to calculate linearity
              for aldrin based on peak height.   It is not permitted within a
              %RSD calculation for an analyte to use calibration factors
              calculated from both peak area and peak height.  For example, it
              is not permitted to calculate the  calibration factor for the low
              point standard for endrin using peak height and calculate the
              midpoint and high point standard calibration factors for endrin
              using peak area.

9.-2.4.5       Calculate the calibration factor for each single component
              pesticide and surrogate over the initial calibration range using
              Equation 2.  The calibration factors for surrogates are calculated
              from the three analyses of Individual Standard A mixture only.

9.2.4.6       Calculate the mean and the %RSD of the  calibration factors for
              each single component pesticide and  surrogate over the initial
              calibration range using Equation 3 and  Equation 4.
                                         D-21/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors  --  Section 3
Calibration and Standardization  (Con't)
              EQ. 2
                     __ m  Peafc area tog Height}  ot the «t«ndmrd
                                     Ibsat Injected
              EQ.  3
                                     er - ^
                                           £ ert
              EQ.  4
                                   *R5D -  —2- * 100
                                            EF
              Where:
               %RSD =      Percent  relative  standard deviation.

   -- ,-s.       SDCP  =      Standard deviation of calibration factors.

   -.: *••       CFi   =       Calibration factor.


               CF  =       Mean calibration  factor.

               n   =       Total number of values  (3).

9.2.4.7       A calibration factor  is  calculated for each peak in a selected set
              of three  to  five  major peaks for each multicomponent analyte using
              Equation  2.

9.2.4.8       Calculate the percent breakdown of DDT, the percent breakdown of
              Endrin, and  the combined breakdown of DDT and Endrin in the
              Performance  Evaluation Mixture  (PEM) using Equations 5, 6,  7,  and
              8.

              EQ.  5

                tftimrt riuti -  P***: *re* (or Peak height)  of compound in PEH
OLC03.2                                 D-22/PEST

-------
                                             Exhibit D Pesticides/Aroclors --  Section  9
                                                Calibration and Standardization (Con't)

             Where:


              CFmp =    The calibration factor for  the compound determined from
                       the midpoint standard in the most recent initial
                       calibration.

             NOTE: If during  the  initial calibration,  linearity was determined
             based on peak area for the compound, then the midpoint CF  must be
             based on peak area.  .If during the initial calibration,  the
             linearity for the compound was determined based on peak height for
             the compound,  then the midpoint CF must be based on peak height.

             EQ. 6

                ^Breakdown DDT -  Amount found (ng>  (DDD+DDE>   .
                                   iaaunc (&a} el DDT injected
             EQ. 7

 ^Breakdown Endrln -  ***""* *<"""> (*>fl) (endrin aldehyde + endrin fcetone)  „
                                   Attoune (&0> at endrin inflected)
             EQ; 8

             Coubined % Breakdown - * Breakdown DDT + % Breakdown Endrin

9.2.4.9      Calculate the percent difference  for each pesticide and surrogate
             in trie PEM using Equations  5  and  9.

             EQ. 9
             Where:

               %D   =      Percent Difference.

               Cnom =     Nominal concentration of each analyte.

               Ccaic  =    Calculated concentration of each  analyte from the
                         analysis of the standard.

9.2.4.10     Calculate the resolution between the analytes in the Resolution
             Check Mixture, PEM and the midpoint  concentrations of Individual
             Standard Mixtures A and B using Equation 10.
                                       D-23/PEST                                OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 9
Calibration and Standardization  (Con't)
              EQ. 10
                                 ^Resolution -  _ * 100
                                                 H
              Where:

               V =   Depth of the valley between the two peaks.  The depth of
                     the.valley is measured along a vertical line from the
                     level of the apex of the shorter peak to the floor of the
                     valley between the two peaks.

               H =   Height of the shorter of the adjacent peaks.

9.2.5      Technical Acceptance  Criteria  for  Initial Calibration

           All  initial  calibration technical  acceptance  criteria  apply
           independently to  each GC column.

9.2.5.1.      The initial calibration sequence must be analyzed according to the
              procedure and in the order listed in Section 9.2.3, at the
              concentrations listed in Section 7.2.3,  and at the frequency
              listed in Section 9.2.2.  The GC/ECD operating conditions
              optimized in Section 9.1 must be followed.

9.2.5.2       The resolution between two adjacent peaks in the Resolution Check
              Mixture must be greater than or equal to 60.0 percent.

9.2.5.3       All single component pesticides and surrogates in both runs of the
              PEM must be greater than or equal to 90.0 percent resolved on each
              column.

9.2.5.4       The absolute RTs of each of the single component pesticides and
              surrogates in both runs of the PEM must be within the RT window
              determined from the three-point initial calibration in Section
              9.2.4.3.

9.2.5.5       The percent difference of the calculated amount (amount found) and
              the nominal amount  (amount added) for each of the single component
              pesticides and surrogates in both of the PEM runs of each GC
              column must be greater than or equal to -25 percent and less than
              or equal to 25 percent using Equation 9.
OLC03.2
D-24/PEST

-------
                                               Exhibit D Pesticides/Aroclors -- Section 9
                                                  Calibration and Standardization (Con't)

9.2.5.6       The  percent  breakdown of  DDT and endrin in each of  the  PEM runs
              must be  less than or equal to 20.0  percent.  The  combined
              breakdown of DDT and endrin must be less than or  equal  to 30.0
              percent.

9.2.5.7       The  %RSD of  the calibration factors for each single component
              target compound must be less than or equal to 20.0  percent,  except
              alpha-BHC and delta-BHC.   The %RSD  of the calibration factors for
              alpha-BHC and delta-BHC must be less than or equal  to 25.0
              percent.   The %RSD of the calibration factors for the two
              surrogates must be less than or equal to 30.0 percent.   Up to two
              single component target compounds (not surrogates)  per column may
              exceed the 20.0 percent limit for %RSD (25.0 percent for alpha-BHC
              and  delta-BHC), but those compounds must have a  %RSD of less than
              or equal to  30.0 percent.

9.2.5.8       The  resolution between any two adjacent peaks in the midpoint
              concentrations of Individual Standard Mixtures A and B in the
              initial  calibration must be greater than or equal to 90.0 percent.

9.2.5.9       All  instrument blanks must meet the technical acceptance criteria
              in Section 12.1.4.5.

9.2.5.10      The  identification of single component pesticides by gas
              chromatographic methods is based primarily on RT data.   The RT of
              the  apex of  a peak can only be verified from an on-scale
              chromatogram.  The identification of multicomponent analytes by
              gas  chromatographic methods is based primarily  on recognition of
              patterns of RTs displayed on a chromatogram. Therefore, the
              following requirements apply to all data presented  for single
              component and multicomponent analytes.

              •    The chromatograms that result  from the analyses of the
                   Resolution Check Mixture,  the  PEM,  and Individual  Standard
                   Mixtures A and B during the initial calibration sequence must
                   display the single component analytes present  in each standard
                   at  greater than 10 percent'of  full scale but less  than 100
                   percent of full scale.

              •    The chromatograms for at least one of the three analyses each
                   of  Individual Standard Mixtures A and B from the initial
                   calibration sequence must display the single component
                   analytes at greater than 50 percent and less than 100 percent
                   of  full scale.

              •    The chromatograms of the standards for the multicomponent
                   analytes analyzed during the initial calibration sequence must
                   display the peaks chosen for identification  of each analyte at
                   greater than 25 percent and less than 100 percent  of full
                   scale.

              •    For all Resolution Check Mixtures, PEMs, Individual Standard
                   Mixtures,  and blanks, the baseline of the  chromatogram must
                   return to below 50 percent of  full scale before the elution
                   time of alpha-BHC, and return  to below 25 percent of full
                   scale after the elution time of alpha-BHC  and  before the
                   elution time of decachlorobiphenyl.

              •    If  a chromatogram is replotted electronically  to meet
                   requirements, the scaling factor used must  be  displayed on  the
                   chromatogram.
                                         D-25/PEST  .                               OLC03.2

-------
   Exhibit D Pesticides/Aroclors -- Section 9
   Calibration and Standardization (Con't)

                 •    If the chromatogram of any standard needs to be replotted
                     electronically to meet these requirements, both the  initial
                     chromatogram and the replotted chromatogram must be  submitted
                     in the data package.

   9.2.6     Corrective Action for Initial Calibration

   9.2.6.1       If  the  technical  acceptance  criteria for the initial  calibration
                 are not met,  inspect  the system for problems.  It  may be necessary
                 to  change  the column, bake out the  detector, clean the injection
                 port, or take other corrective actions  to achieve  the technical
                 acceptance criteria.

   9.2.6.2       Contamination should  be  suspected as a  cause if  the detector
                 cannot  achieve acceptable linearity using this method.   In the
                 case of low level contamination,  baking out the  detector at
                 elevated temperature  (350°C)  should be  sufficient  to achieve
                 acceptable performance.   In  the case of heavy contamination,
                 passing hydrogen  through the detector for 1-2 hours at elevated
                 temperature may correct  the  problem.   In the case  of severe
                 contamination,  the detector  may require servicing  by the BCD
                 manufacturer.  DO NOT OPEN THE DETECTOR.   THE BCD  CONTAINS
                 RADIOCHEMICAL SOURCES.

   9.2.6.3       If  a laboratory cleans out a detector using elevated temperature,
     '-  '         the BCD electronics must be  turned  off  during the  bake out
                 procedure.

 '  9.2.6\4       After bake out or hydrogen reduction, the detector must be
-'.  ••'.'-•  ''"        recalibrated using the initial calibration sequence.
    . ,, .   ->•
   9.2.6.5       Initial calibration technical acceptance criteria  MUST be met
 •v; -*:"?' "..        before  any samples (including Laboratory Control Samples, matrix
 ••''•'•"''•*  "';        spike,  and matrix spike  duplicate,  if required)  or required blanks
'¥'"'•"-"  "'        (method/sulfur clean-up)  are analyzed.   Any samples or required
 •:  •'!;'  '         blanks  analyzed when  the initial calibration technical acceptance
-';*•'  ••?-,- ""         criteria have not been met will require re-analysis at no
                 additional cost to USEPA.

  .9.3    Calibration Verification

   9.3.1     Summary of Calibration Verification

             Three  types of analyses are used to verify the  calibration and
             evaluate instrument  performance.  The  analyses  of instrument blanks,
            • PEMs,  and  the mid point concentration  of Individual Standard Mixtures
             A and  B constitute the calibration verification.  Sample  (including
             Laboratory Control Sample and matrix spike and  matrix  spike
             duplicate, if required) and required blank (method/sulfur  clean-up)
             data are not  acceptable unless bracketed by acceptable analyses of
             instrument blanks, PEM, and both Individual Standard Mixtures A and
             B.

   9.3.2     Frequency  of  Calibration Verification

   9.3.2.1       An instrument blank and  the  PEM must bracket one end of a 12-hour
                 period  during which sample and required blank data are collected,
                 and a second instrument  blank and the mid point  concentration of
                 Individual Standard Mixtures A and  B must bracket  the other end of
                 the 12-hour period.
   OLC03.2                                 D-26/PEST

-------
                                               Exhibit D Pesticides/Aroclors -- Section 9
                                                  Calibration and Standardization  (Con't)

9.3.2.2       For the 12-hour period immediately following the  initial
              calibration sequence,  the instrument  blank and the PEM that are
              the last two steps in the initial calibration sequence bracket the
              front end of that 12-hour period.  The injection  of the instrument
              blank starts the beginning of the 12-hour period  (Section
              10.2.2.1).   Samples [including Laboratory Control Samples (LCSs)
              and matrix spike and matrix spike duplicate (MS/MSD),  if  required]
              and required blanks (method/sulfur clean-up)  may  be injected for
              12 hours from the injection of the instrument blank.   The first
              three injections immediately after that 12-hour period must be an
              instrument blank. Individual Standard Mixture A,  and Individual
              Standard Mixture B.  The instrument blank must be analyzed first,
              before either standard.  The Individual Standard  Mixtures may be
              analyzed in either order (A,B or B,A).

9.3.2.3       The analyses of the instrument blank  and Individual Standard
              Mixtures A and B immediately following one 12-hour period may be
              used to begin the subsequent 12-hour  period, provided that they
              meet the technical acceptance criteria in Section 9.3.5.   In that
              instance, the subsequent 12-hour period must be bracketed by the
              acceptable analyses of an instrument  blank and a  PEM,  in that
              order.  Those two analyses may in turn be used to bracket the
              front end of yet another 12-hour period.  This progression may
              continue every 12 hours until such time as any of the instrument
              blanks, PEMs, or Individual Standard  Mixtures fails to meet the
              technical acceptance criteria in Section 9.3.5.  The 12-hour time
              period begins with the injection of the instrument blank.

9.3.2.4       Standards  (PEM or Individual Standard Mixtures),  samples and
              required blanks may be injected  for 12 hours from the time of
              injection of the instrument blank.

9.3.2:5       If more than 12 hours have elapsed since the injection of the
              instrument blank that bracketed.a previous 12-hour period, an
             - acceptable instrument blank and  PEM,must be. analyzed in order to
             . start a new sequence. . This requirement applies even if no
              analyses were performed since that standard was injected.

9.3.2.6       The requirements for running the instrument blanks, PEM,  and
              Individual Standard Mixtures A and B are waived when no samples
              (including Laboratory  Control Samples, requested matrix spike and
              matrix spike duplicate) dilutions, re-analyses, required blanks
              (method/sulfur clean-up), and multicomponent analytes for the 72-
              hour  confirmation requirement are analyzed during that 12-hour
              period.  To resume analysis, using the existing initial
              calibration, the Contractor must first analyze an instrument blank
              and PEM which meet the technical acceptance criteria.

9.3.2.7       If the entire  12-hour  period is  not  required for the analyses of
              all samples and  blanks to be reported and all data collection is
              to be stopped, the sequence must be  ended with either the
              instrument blank/PEM combination or  the  instrument
              blank/Individual Standard Mixtures A and B combination, whichever
              was due  to be performed  at  the end of the 12-hour period.

9.3.3     Procedure for Calibration Verification

9.3.3.1       All Standards  and  blanks must warm to ambient temperature prior  to
              analysis.
                                         D-27/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 9
Calibration and Standardization  (Con't)

9.3.3.2       Analyze the instrument blank, PEM, and the mid point concentration
              of Individual Standard Mixtures A and B at the required
              frequencies (Section 9.3.2).

9.3.4     Calculations  for  Calibration Verification

9.3.4.1       For each analysis of the PEM used to demonstrate calibration
              verification,  calculate the percent difference between the amount
              of each analyte (including the surrogates) found in the PEM and
              the nominal amount, using Equations 5 and 9.

9.3.4.2       For each analysis of the PEM used to demonstrate calibration
              verification,  calculate the percent breakdown of Endrin and DDT,
              and the combined breakdown, using Equations 5, 6, 7, and 8.

9.3.4.3       For each analysis of the mid point concentration of Individual
              Standard Mixtures A and B used to demonstrate calibration
              verification,  calculate the percent difference between the amount
              of each analyte (including the surrogates) found in the standard
              mixture and the nominal amount, using Equations 5 and 9.  Do not
              attempt to calculate the breakdown of Endrin and DDT in the
              Individual Standard Mixtures, as these standards contain the
              breakdown products as well as the parent compounds.

9.3.5     Technical  Acceptance  Criteria For Calibration Verification

    ..- ..   All  calibration verification technical  acceptance criteria apply
          independently to  each GC column.

9.3:5.'l       The PEMs,  Individual Standard Mixtures, and instrument blanks must
    ~ :       be analyzed at the' required frequency  (Section 9.3.2), on a GC/ECD
              system that has met the initial calibration technical acceptance
    ^U       criteria.

9.3V5V2       All single component pesticides and surrogates in the PEMs used to
    •"'• -.       demonstrate calibration verification must be greater than or equal
    ~?K:::       to 90.0 percent resolved.  The resolution between any two adjacent
              peaks in the midpoint concentrations of Individual Standard
              Mixtures A and B used to demonstrate calibration verification must
              be greater than or equal to 90.0 percent.

9.3.5.3       The absolute RT for each of the single component pesticides and
              surrogates in the PEMs and mid point concentration of the
              Individual Standard Mixtures used to demonstrate calibration
              verification must be within the RT windows determined from the
              three-point initial calibration in Section 9.2.4:3.

9.3.5.4       The percent difference of the calculated amount  (amount found) and
              the nominal amount (amount added) for each of the single component
              pesticides and surrogates in the PEM and midpoint concentration of
              the Individual Standard Mixtures used to demonstrate calibration
              verification must be greater than or equal to -25.0 percent and
              less than or equal to 25.0 percent.

9.3.5.5       The percent breakdown of 4,4"-DDT in the PEM must be less than or
              equal to 20.0 percent on each column.  The percent breakdown of
              Endrin in the PEM must be less than or equal to 20.0 percent on
              each column.  The combined breakdown of DDT and endrin must be
              less than or equal to 30.0 percent on each column.

9.3.5.6       All instrument blanks must meet the technical acceptance criteria
            •  in Section 12.1.4.5.


OLC03.2                                  D-28/PEST

-------
                                               Exhibit D Pesticides/Aroclors >•- Section 9
                                                  Calibration and Standardization (Con't)

9.3.5.7       The  identification of single component  pesticides  by gas
              chromatographic methods is based primarily on RT data.   Since the
              RT of  the apex of a peak can only be verified from an on-scale
              chromatogram,  the following requirements must be met for
              calibration verification to be acceptable:

              •   The  chromatograms  that  result from the analyses of  the PEM and
                  the  Individual  Standard Mixtures must display the single
                  component  analytes present  in each standard at greater than  10
                  percent of full scale but less,  than 100  percent of  full  scale;

              •   For  any PEM,  Individual Standard Mixture, or  blank,  the
                  baseline  of the chromatogram must  return to below 50 percent
                  of full scale before  the  elution time of alpha-BHC,  and  return
                  to below  25 percent of  full scale  after  the elution time  of
                  alpha-BHC and before  the  elution time of decachlorobiphenyl;

              •   If a chromatogram is  replotted  electronically to meet these
                  requirements,  the scaling factor used must  be displayed  on the
                  chromatogram;  and

              •   If the chromatogram of  any  standard or blank  needs  to be
                  replotted electronically to meet these requirements, both the
                  initial chromatogram and the replotted chromatogram(s) must  be
                  submitted in the data package.

9.3.6      Corrective Action for Calibration Verification

9.3.6.1       If the technical acceptance criteria for the calibration
              verification are not met, inspect the system for problems and take
              corrective action to achieve the technical acceptance criteria.

9.3.6.2       Major corrective actions such as replacing the GC  column or baking
              out the detector will  require that a new initial calibration be
              performed that  meets the technical acceptance criteria  in Section
              9.2.5.

9.3.6.3       Minor corrective actions may not require performing a new initial
              calibration, provided  that  a new analysis of the  standard (PEM or
              Individual Mixtures) that originally failed  the criteria and an
              associated instrument  blank immediately after the  corrective
              action do meet  all  the technical acceptance  criteria.

9.3.6.4       If a PEM or Individual Standard Mixture does not meet the
              technical acceptance criteria listed above,  it must be  reinjected
              immediately.   If  the second injection  of  the PEM  or Individual
              Standard Mixture meets the  criteria, sample  analysis may continue.
              If the second  injection does not meet  the criteria, all  data
              collection must be  stopped.  Appropriate  corrective action must  be
              taken, and a new initial calibration sequence must be run before
              more sample data are collected.

9.3.6.5       If an instrument  blank does not meet the  technical acceptance
              criteria  listed in  Section  12.1.4.5, all  data collection must be
              stopped.  Appropriate  corrective action must be taken to clean out
              the system, and an  acceptable instrument  blank must be  analyzed
              before more sample  data are collected.

9.3.6.6       Analysts  are  reminded  that  running  an  instrument  blank  and a  PEM
              or  Individual  Standard Mixtures once every 12 hours are the
              minimum  contract requirements.   Late eluting peaks may  carry  over
               from one injection  to  the next  if highly complex  samples are


                                         D-29/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 9
Calibration and Standardization (Con't)

              analyzed or if the GC conditions are unstable.  Such carryover is
              unacceptable.   Therefore, it may be necessary to run instrument
              blanks and standards more often to avoid discarding data.

9.3.6.7       If a successful instrument blank and PEM cannot be run after an
              interruption in analysis (Section 9.3.2.6),  an acceptable  initial
              calibration must be run before sample data may be collected.  All
              acceptable sample (including Laboratory Control Samples, requested
              matrix spike and matrix spike duplicate)  and required blank
              (method/sulfur clean-up) analyses must be preceded and followed by
              acceptable standards and instrument blanks,  as described in
              Section 9.3.2.

9.3.6.8       Calibration verification technical acceptance criteria must be met
              before, any samples (including the Laboratory Control Sample,
              requested matrix spike and matrix spike duplicate) and required
              blanks (method/sulfur clean-up)  are reported.  Any samples and
              required blanks associated with a calibration verification which
              did not meet the technical acceptance criteria will require re-
              analysis at no additional cost to USEPA.
OLC03.2                                 D-30/PEST

-------
                                              Exhibit D Pesticides/Aroclors -- Section 10
                                                                                Procedure

10.0   PROCEDURE

10.1   Sample Preparation

10.1.1    This method is designed for analysis of water samples that contain
          low concentrations of the pesticides and Aroclors listed in Exhibit
          C.  The majority of the samples are expected to come from drinking
          water and well/ground water type sources around Superfund sites.  If,
          upon inspection of a sample, the Contractor suspects that the sample
          is not amenable to this method, contact Sample Management Office
          .(SMO).  SMO will contact the Region for instructions.

10.1.2    If insufficient sample volume  (less than 90 percent of the required
          amount) is received to perform the analyses, the Contractor shall
          contact SMO to apprise them of the problem.  SMO will contact the
          Region for instructions.  The  Region will either require that no
          sample analyses be performed or will require that a reduced volume be
          used for the  sample analysis.  No other changes in the analyses will
          be permitted.  The Contractor  shall document the Region's decision in
          the Sample Delivery Group  (SDG) Narrative.

10.1.3    Extraction of Samples

          Water samples may be extracted by either separatory funnel procedure
          or a  continuous liquid-liquid  extraction procedure.  If an emulsion
          prevents acceptable solvent recovery with the separatory funnel
          procedure, continuous  liquid-liquid extraction must be employed.
          Allow the samples to come  to ambient temperature  (approximately one
          hour).

10.1.3.1      Separatory Funnel Extraction

10.1.3.1.1       Measure out each 1 L sample aliquot  in a separate graduated
                 cylinder.  Measure and.record the pH of the sample with wide
                 range  pH paper  and adjust the pH to  between 5 and 9 with 10 N
                 sodium hydroxide or concentrated sulfuric acid, if required.
                 Samples requiring pH adjustment must be noted in the SDG
                 narrative.  Place the sample aliquot into a 2 L separatory
                 funnel.

10.1.3.1.2       Using  a syringe or  a volumetric pipet add 200 microliter  (uL)
                 of the surrogate solution  (Section 7.2.3.1) to all water
                 samples.

10.1.3.1.3       Rinse  the graduated cylinder with 30 milliliters  (mL) of
                 methylene chloride  and  transfer the  rinsate to the separatory
                 funnel. If  the  sample container is empty, rinse the container
                 with  30 mL  of methylene chloride and add the rinsate to the
                 separatory  funnel.   If  the sample container is not rinsed, then
                 add another  30  mL of methylene chloride to the separatory
                 funnel and  extract  the  sample by shaking the funnel for two
                 minutes,  with periodic  venting to release excess pressure.

                 NOTE:  The  total volume  of  solvent used for extraction  is  60 mL.
                 Allow the  organic  layer to separate  from the water phase  for a
                 minimum of  10 minutes.   If the emulsion interface between
                 layers is more  than one-third the volume of the solvent  layer,
                 the analyst  must  employ mechanical  techniques  to complete the
                 phase separation.   The  optimum technique depends upon  the
                 sample, and may include stirring, filtration through glass
                 wool,  centrifugation or other physical means.  Drain the
                 methylene chloride  into a  250 mL Erlenmeyer  flask.


                                         D-31/PEST                                  OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure (Con't}

10.1.3.1.4       Add  a  second 60 mL volume of methylene  chloride to the
                 separatory  funnel and repeat the  extraction procedure a second
                 time,  combining the extracts in the  Erlenmeyer flask.   Perform
                 a  third extraction in the same  manner.

10.1.3.2      Continuous Liquid-Liquid Extraction

10.1.3.2.1       Continuous  Liquid-Liquid Extraction  Without Hydrophobic
                 Membrane

10.1.3.2.1.1         Follow manufacturer's instructions for set-up.

10.1.3.2.1.2         Add methylene chloride to the bottom of the extractor and
                     fill it to a depth of at least one inch above the bottom
                     sidearm.

10.1.3.2.1.3         Measure out each 1 L sample aliquot in a separate, clean
                     graduated cylinder;  transfer the aliquot to the continuous
                     extractor.   Measure the pH of the sample with wide range pH
                     paper or pH meter and record pH.   Adjust the pH to between 5
                     and 9 with 10 N sodium hydroxide or concentrated sulfuric
                     acid, as required.   Samples requiring the pH adjustment must
                     be noted in the SDG Narrative.

                     NOTE: With some samples, it may be necessary to place a
      s'.              layer of glass wool between the methylene chloride and the
     •••..••             water layer in the extractor to prevent precipitation of
      -«              suspended solids into the methylene chloride during
                     extraction.

10 .1.3;;,2 .1.4         Using a syringe or volumetric pipet, add 200 uL of the
                     surrogate standard spiking solution (Section 7.2.3.1) into
      •/?'•>             the sample and mix well.
      i'3'~-'
10.1.3V2.1.5         Rinse the graduated cylinder with 50 mL of methylene
      „'-              chloride and transfer the rinsate to the continuous
      --•              extractor.   If the sample container is empty, rinse the
                     container with 50 mL of methylene chloride and add the
                     rinsate to the continuous extractor.

10.1.3.2.1.6         Add sufficient methylene chloride to the continuous
                     extractor to ensure proper solvent cycling during operation.
                     Adjust the drip rate to 5 to 15 mL/minute  (recommended);
                     optimize the extraction drip rate.  Extract for a minimum of
                     18 hours.

                     NOTE 1: When a minimum drip rate of 10-15 tnLs/minute is
                     maintained throughout the extraction,  the extraction time
                     may be reduced to a minimum of 12 hours.  Allow to cool,
                     then detach the distillation flask.  Proceed to Section
                     10.1.4.

                     NOTE 2: Some continuous liquid-liquid extractors are also
                     capable of concentrating the extract within the extraction
                     set-up.  Follow the manufacturer's instructions for
                     concentration when using this type of extractor.
OLC03.2                                  D-32/PEST

-------
                                              Exhibit D Pesticides/Aroclors -- Section 10
                                                                        Procedure  (Con't)

10.1.3.2.2       Continuous Liquid-Liquid Extraction With Hydrophobic Membrane

10.1.3.2.2.1         Follow the manufacturer's instructions  for set-up.

10.1.3.2.2.2         Measure out each 1 L sample aliquot in  a separate,  clean
                     graduated cylinder.  If the sample container is empty,  rinse
                     the container with 50 mL of methylene chloride and add the
                     rinsate to the continuous extractor.  If the sample was not
                     received in a 1 L container, add 50 mL  of methylene chloride
                     to the continuous extractor.  Slowly transfer the aliquot to
                     the continuous extractor.  Measure the  pH of the sample with
                     wide range pH paper or pH meter and record pH.  Adjust the
                     pH to between 5 and 9 with 10 N sodium  hydroxide or
                     concentrated sulfuric acid, as required.  Samples requiring
                     the pH adjustment must be noted in the  SDG Narrative.

10.1.3.2.2.3         Using a syringe or volumetric pipet,  add 200 uL of the
                     surrogate standard spiking solution (Section 7.2.3.1) into
                     the sample and mix well.

10.1.3.2.2.4        .Rinse the graduated cylinder with 50 mL of methylene
                     chloride and transfer the rinsate to the continuous
                     extractor.

10.1.3.2.2.5         Add sufficient methylene chloride to the continuous
                     extractor to ensure proper solvent cycling during operation.
                     Adjust the drip rate to 15 mL/minute (recommended); optimize
                     the extraction drip rate.  Extract for a minimum of 6 hours.

                     NOTE 1: Due to the smaller volume of solvent used during the
                     extraction process, some .sample matrices  (e.g., oily
                     samples, samples containing a high concentration of
                     surfactants) may create, an .emulsion which will consume the
                     solvent volume, preventing the efficient extraction of the
                     sample.'  When this-occurs, add additional solvent to ensure
                     efficient extraction of the sample, and extend the
                     extraction time for a minimum of 6 hours.  If the sample
                     matrix prevents the free flow of solvent through the
                     membrane, then the non-hydrophobic membrane continuous
                     liquid-liquid type extractor must be used.

                     Allow to cool, then detach the distillation flask.  Proceed
                     to Section 10.1.4.

                     NOTE 2: Some continuous liquid-liquid extractors are also
                     capable of concentrating the extract within the extraction
                     set-up.  Follow the manufacturer's instructions for
                     concentration when using this type of extractor.  Using the
                     hydrophobic membrane,  it may not be necessary to dry the
                     extract with sodium sulfate.

                     NOTE 3: If low surrogate recoveries occur, assure 1) the
                     apparatus was properly assembled to prevent leaks,- 2)  the
                     drip rate/solvent  cycling  was optimized; and  3) there  was
                     proper cooling for condensation of solvent.   Document  the
                     problem and the  corrective action.

                     NOTE 4: Alternate  continuous liquid-liquid extractor types
                     that meet  the  requirements of the  SOW may also be used.   If
                     using  alternate  extractors or design types, follow the
                     manufacturer's instructions  for set-up.  Optimize the
                     extraction procedure.


                                         D-33/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure (Con't)

10.1.4    Extract Drying and Concentration

10.1.4.1      Assemble a kuderna-Danish (K-D)  concentrator by attaching a 10 mL
              concentrator tube to a 500 mL evaporative flask.  Other
              concentration devices or techniques may be used in place of the
              K-D,  if equivalency is demonstrated for all the target pesticides
              and Aroclors listed in Exhibit C.

10.1.4.1      Pour the extract  through a drying column containing about 10
              centimeters (cm)  of anhydrous granular sodium sulfate and collect
              the extract in the K-D concentrator.  Rinse the Erlenmeyer flask
              and the sodium sulfate with at least two additional 20 to 30 mL
              portions of methylene chloride to complete the quantitative
              transfer.

10.1.4.3      Add one or two clean boiling chips to the evaporative flask and
              attach a three-ball Snyder column. Pre-wet the Snyder column by
              adding about 1 mL of methylene chloride to the top of the column.
              Place the K-D apparatus on a hot water bath (60-90°C recommended)
              so that the concentrator tube is partially immersed in the hot
              water and the entire lower rounded surface of the flask is bathed
              with hot vapor.   Adjust the vertical position of the apparatus and
              the water temperature as required to complete the concentration in
              15 to 30 minutes.  At the proper rate of distillation, the balls
              of the column will actively chatter, but the chambers will  not
      *.'       flood with condensed solvent.  When the apparent volume of liquid
      ?       reaches 3 to 5 mL, remove the K-D apparatus.  Allow it to drain
              and cool for at least 10 minutes.  DO NOT ALLOW THE EVAPORATOR TO
              GO DRY.  Proceed with the solvent exchange to hexane.
      "^:
10.1.4>4      Solvent Exchange to Hexane
      X*
10.1.47-4.1       Momentarily remove  the three-ball  Snyder column,  add  50 mL  of
     -Kit          hexane and a new boiling chip,  and reattach  the  Snyder column.
      "S-          Pre-wet the column  by adding  about 1 mL of hexane  to  the top.
      '•••;,          Concentrate the solvent extract as before.   When the  apparent
      :**          volume of  liquid reaches  1 mL,  remove  the K-D apparatus and
                 allow it to drain and cool for at  least 10 minutes.   DO NOT
                 ALLOW THE  EVAPORATOR TO GO DRY.

10.1.4.4.2       Remove the Snyder column; using 1  to 2  mL of hexane,  rinse  the
                 flask and  its  lower joint into the concentrator  tube.

10.1.4.4.3       Use the micro  Snyder column or the nitrogen  blowdown  technique
                 to  concentrate the  hexane extract  to 2.0 mL.

10.1.4.5      Final Concentration of Extract

              Two different techniques are permitted to concentrate the extract
              to 2.0 mL.   They are the micro Snyder column and nitrogen
              evaporation techniques.

10.1.4.5.1       Micro Snyder Column Technique

                 Add another one or  two clean  boiling chips to the  concentrator
                 tube and attach a two-ball micro Snyder column.  Pre-wet the
                 Snyder column  by adding about 0.5  mL of hexane to  the top of
                 the column.  Place  the K-D apparatus in a hot water bath  (80°C
                 to  90°C recommended) so that  the concentrator tube is partially
                 immersed in the hot water.  Adjust the  vertical  position of the
                 apparatus  and  the water temperature as  required  to complete the
                 concentration  in 5  to 10 minutes.   At  the proper rate of


OLC03.2                                 D-34/PEST

-------
                                              Exhibit D Pesticides/Aroclors -- Section 10
                                                                        Procedure (Con't)

                 distillation the balls of the column will actively chatter but
                 the chambers will not flood with condensed solvent. When the
                 apparent volume of liquid reaches about 0.5 inL, remove the K-D
                 apparatus from the water bath and allow it to drain for at
                 least 10 minutes while cooling.  Remove the Snyder column and
                 rinse its flask and lower joint into the concentrator tube with
                 0.2 mL of hexane.  Adjust the final volume with hexane to 2.0
                 mL.

10.1.4.5.2       Nitrogen Evaporation Technique (taken from ASTM Method D 3086).

10.1.4.5.2.1         Place the  concentrator  tube with an  open  micro Snyder  column
                     attached in a warm water  bath (30°C  to 35°C recommended)  and
                     evaporate  the solvent volume  to just below 1 or 2  mL by
                     blowing a  gentle  stream of  clean,  dry nitrogen filtered
                     through a  column  of activated carbon above the solvent.
                     Adjust the final  volume with  hexane  to 1.0 mL (Florisil
                     cartridge  check)  or 2.0 mL  (sample extract).

10.1.4.5.2.2         CAUTION: Gas lines from the gas source to the evaporation
                     apparatus  must be stainless steel, copper, or PTFE tubing.
                     Plastic tubing must not be  used between the carbon trap and
                     the sample as it  may introduce interferences.   The internal
                     wall of new tubing must be  rinsed several times with hexane
                     and then dried prior to use.

10.1.4.5.2.3         During evaporation, the tube  solvent level must be kept
                     below the  water level of  the  bath.   DO NOT ALLOW THE EXTRACT
                     TO GO TO DRYNESS.

10.1.4.6      Proceed to Section 10.1.5 for extract cleanup.  Otherwise,
              transfer the extract to a PTFE-lined screw-cap bottle and label
              the bottle.  Store at 4°C  (±2°C) but not greater than 6°C.

10.1.5     Extract  Cleanup

10.1.5.1      The two cleanup procedures specified in this method are Florisil
              cartridge and  sulfur cleanup.   Florisil cartridge cleanup is
              required for all extracts.  Sulfur cleanup must be performed on
              all extracts containing sulfur at levels that interfere with Gas
              Chromatograph/Electron Capture Detector  (GC/ECD) analysis.  Sulfur
              contamination  in a sample analysis is unacceptable.  Method blanks
              must be subjected to the same cleanup procedures as the samples.

10.1.5.2      Florisil Cleanup

10.1.5.2.1        Introduction

                  Florisil cartridge cleanup significantly reduces matrix
                  interference  caused  by polar compounds  and is required for all
                  extracts.  The same  volume of  the concentrated extract taken
                  for Florisil  cleanup must  be maintained after Florisil cleanup
                  (2 mL) .

10.1.5.2.2        Florisil Cartridge Performance Check

10.1.5.2.2.1         Summary of Florisil Cartridge Performance Check

                     Every lot number of Florisil cartridges must be tested
                     before  they'are used for sample cleanup.
                                         D-35/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure ( Con ' t )

10.1.5.2.2.2         Frequency of Florisil Cartridge Performance Check

                     Cartridge performance check must be conducted at least once
                     on each lot of cartridges used for sample cleanup or every 6
                     months whichever is most frequent.

10.1.5.2.2.3         Procedure for Florisil Cartridge Performance Check

                     Add 0.5 tnL of 2,4,5-trichlorophenol solution (0.1 ug/mL in
                     acetone, Section 7.2.3.7) and 0.5 mL of Standard Mixture A,
                     midpoint concentration.  Section 7.2.3.5)  to 4 mL of hexane.
                     Reduce the volume to 0.5 mL using nitrogen (Section
                     10.1.4.5.2).  Place the mixture onto the top of a washed
                     Florisil cartridge, and elute it with 9 mL of hexane/acetone
                     [(90:10) (V/V) ] .   Use two additional 1 mL hexane rinses to
                     ensure quantitative transfer of standard from the cartridge .
                     Concentrate to a final volume of 1 mL and analyze the
                     solution by GC/ECD using at least one of the GC columns
                     specified for sample analysis .   Determine the recovery of
                     each arialyte for evaluation and reporting purposes.
                     Calculate the percent recovery using Equation 11.

                     EQ. 11

                              Percent Rtcovfcry - — » 100
                     Where :

      .-!_..             Qd =    Quantity determined by analysis.

      ,^T             Qa =    Quantity added.

10.1 .5. ,2 .2 .4         Technical Acceptance Criteria for Florisil Cartridge
                     Performance Check
      ~i>:- '
10.1.5.2.2.4.1          The cartridge performance check solution must be analyzed
                        on a  GC/ECD meeting the  initial calibration and
                        calibration verification technical acceptance criteria.

10.1.5.2.2.4.2          The lot of Florisil cartridges is acceptable if all
                        pesticides are recovered at 80 to 120 percent, if the
                        recovery of trichlorophenol is less  than 5 percent, and
                        if no peaks interfering  with the target analytes are
                        detected .

10.1.5.2.2.5         Corrective Action for Florisil Cartridge Performance Check

                     Any lot of Florisil cartridges that does not meet the
                     criteria above must be discarded and a new lot,  meeting
                     criteria, used for sample cleanup.

10.1.5.2.3        Sample Cleanup by Florisil Cartridge

                  The required Florisil cartridge size is a 1 g cartridge and the
                  final volume of the extract  after Florisil  cleanup is 2 mL.
OLC03.2                                 D-36/PEST

-------
                                              Exhibit D Pesticides/Aroclors -- Section 10
                                                                        Procedure  (Con't)

10.1.5.2.3.1         Frequency of Sample Cleanup by Florisil Cartridge

                     All sample extracts (including Laboratory Control Samples
                     and requested matrix spike  and matrix spike duplicate)  and
                     method blank extracts are required to be cleaned up by the
                     Florisil cartridge technique.

10.1.5.2.3.2         Procedure for Sample Cleanup by Florisil Cartridge

10.1.5.2.3.2.1          Attach the vacuum manifold to a water aspirator or to a-
                        vacuum pump with a trap installed between the manifold
                        and  the vacuum  source.  Adjust the vacuum pressure in the
                        manifold  to between 5 and 10 pounds of vacuum.

10.1.5.2.3.2.2          Place one Florisil cartridge into the vacuum manifold for
                        each sample extract.

10.1.5.2.3.2.3          Prior to  cleanup of samples, the cartridges must be
                        washed with hexane/acetone  (90:10).  This is  accomplished
                        by placing the  cartridge on the vacuum manifold, by
                        pulling  a vacuum, and by passing at  least 5 mL of the
                       . hexane/acetone  solution through the  cartridge.  While the
                        cartridges are  being  washed, adjust  the  vacuum applied to
                        each cartridge  so that the  flow rate through  each
                        cartridge is approximately  equal.  DO NOT ALLOW THE
                        CARTRIDGES TO GO DRY  AFTER  THEY HAVE BEEN WASHED.

10.1.5.2.3.2.4          After  the cartridges  on the manifold are washed, the
                        vacuum is released, and a rack containing labeled  10 mL
                        volumetric flasks is  placed inside  the  manifold.   Care
                        must be  taken  to  ensure that the  solvent line from each
                         cartridge is placed inside-of the  appropriate volumetric
                         flask  as the manifold top  is replaced.

10.1.5.2.3.2.5          After  the volumetric  flasks are  in place, the vacuum  to
                         the manifold is restored,  and a volume  of extract  equal
                         to the required final volume from each  sample and  method
                         blank extract  is  transferred to  the top frit  of  the
                         appropriate Florisil  cartridge.   This must  equal the
                         final  volume after Florisil cleanup.

10.1.5.2.3.2.6           Because the volumes marked on  concentrator  tubes are  not
                         necessarily accurate  at  the 1 mL  level,  the use  of a
                         syringe or a volumetric  pipet  is  required to transfer the
                         extract to the cleanup  cartridge.

10.1.5.2.3.2.7           The pesticides/Aroclors in the  extract  concentrates are
                         then eluted through the column with 8  mL of
                       • hexane/acetone (90:10)  and are  collected into the  10  mL
                         volumetric flasks held  in the  rack inside the vacuum
                         manifold.

10.1.5.2.3.2.8           Transfer the eluate in  each volumetric  flask to a  clean
                         centrifuge tube or 10 mL vial.   Use two additional 1  mL
                         hexane rinses  to ensure quantitative transfer of the
                         cartridge eluate.

10.1.5.2.3.2.9           Adjust the extract to 2 mL aliquot volume as was taken
                         for cleanup using either nitrogen blowdown  or a micro
                         Snyder column   (Section  10.1.4.5).   Measure  the final
                         volume with  a  syringe or by transferring the extract  to  a
                         volumetric flask.


                                         D-37/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors  -- Section 10
Procedure (Con't)

10.1.5.2.3.2.10         If sulfur cleanup is to be performed, proceed  to  Section
                        10.1.5.3.3.  Otherwise, transfer the  sample  to a  GC vial
                        and  label the vial.  The extract is ready  for  GC/ECD
                        analysis.

10.1.5.3      Sulfur Cleanup

10.1.5.3.1        Introduction

10.1.5.3.1.1         Sulfur contamination will cause a rise in the baseline of a
                     chromatogram and may interfere with the analyses of the
                     later eluting pesticides.   If crystals of sulfur are evident
                     or if the presence of sulfur is suspected,  sulfur removal
                     must be performed.   Interference which is due to sulfur is
                     not acceptable.  Sulfur can be removed by one of two
                     methods, according to laboratory preference.  If the sulfur
                     concentration is such that crystallization occurs in the
                     concentrated extract,  centrifuge the extract to settle the
                     crystals,  and remove the sample extract with a disposable
                     pipette, leaving the excess sulfur in the centrifuge tube.
                     Transfer the extract to a clean centrifuge tube or clean
                     concentrator tube before proceeding with further sulfur
                     cleanup.

10.1.5.3.1.2         If only part of a set of samples require sulfur cleanup,
                     then, a sulfur cleanup blank is required for that part of
                     the set (Section 12.1.3.2).

•10.1.5.3'.2        Frequency of  Sulfur Cleanup

       -'          Sulfur removal  is  required for all  sample extracts  that  contain
•'•'57     '<•"          sulfur.

10.1.5.3.3        Procedure for Sulfur Cleanup

•1-0-. 1.5.3-.3.1         Mercury Technique

                     Add one to three drops of mercury to each hexane extract in
                     a clean vial.  Tighten the top on the vial and agitate the
                     sample for 30 seconds.  Filter or centrifuge the extract.
                     Pipet the extract to another vial and leave all solid
                     precipitate and liquid mercury.   If the mercury appears
                     shiny, proceed to Section 10.2 and analyze the extract.  If
                     the mercury turns black,  repeat sulfur removal as necessary.
                     The extract transferred to the vial still represents the 2.0
                     mli final volume.  CAUTION: Waste  containing mercury should
                     be segregated and disposed of properly.

                     NOTE: Mercury is a highly toxic metal and therefore must be
                     used with great care.  Prior to using mercury, it is
                     recommended that the analyst become acquainted with proper
                     handling and cleanup techniques associated with this metal.

10.1.5.3.3.2         Copper Technique

                     Add approximately 2 grams  (g)  of cleaned copper powder to
                     the extract in the centrifuge or concentrator tube (2 g will
                     fill the tube to about the 0.5 mL mark).   Mix the copper and
                     extract for at least 1 minute on a mechanical shaker.
                     Separate the extract from the copper powder by drawing off
                     the extract with a disposable pipet, and transfer the
                     extract to a clean vial.   The extract transferred to the


OLC03.2                                  D-38/PEST

-------
                                              Exhibit D Pesticides/Aroclors -- Section 10
                                                                        Procedure (Con't)

                    vial  still represents the 2.0 mL final volume.  The
                    separation of the extract from  the  copper powder  is
                    necessary to prevent degradation of the pesticides.   If the
                    copper   appears bright, proceed to  Section  10.2 and analyze
                    the extracts.  If the copper changes color,  repeat the
                    sulfur  removal procedure as necessary.

10.2   GC Analysis

10.2.1    Introduction

10.2.1.1      Before samples (including  Laboratory  Control Samples and requested
              matrix spike and matrix  spike  duplicate) and required blanks
              (method/sulfur clean-up) can be  analyzed,  the instrument must  meet
              the initial  calibration  and calibration verification technical
              acceptance criteria.  All  sample extracts, required blanks,  and
              calibration standards must be  analyzed under the  same instrumental
              conditions.   All sample  extracts,  required blank  extracts,  and
              standard/spiking solutions must  be allowed to warm to ambient
              temperature (approximately 1 hour)  before preparation/analysis.
              Sample analysis on both  GC columns is required for all  samples and
              blanks.

10.2.1.2      Set up the GC/ECD system per the requirements in  Section 9.1.
              Unless ambient temperature on-column injection is used  (Section
              9.1), the injector must.be heated to at least 200°C. The
              optimized gas chromatographic  conditions from Section 9.1 must be
              used.

10.2.2    Procedure for Sample Analysis by  GC/ECD

          The injection must be made  on-column by using either  automatic  or
          manual injection.   If  autoinjectors are used, 1.0 uL  injection
          volumes may be  used.   Manual  injections must  use at  least 2.0 uL
          injection volumes. The same  injection volume must be used  for  all
          standards,  samples, and'blanks associated:with the same initial
          calibration.   If  a single injection is used for two  GC columns
          attached to a  single  injection port,  it may be necessary to use an
          injection volume greater than 2 uL.   However, the same injection
          volume must be  used for all analyses.

10.2.2.1      Analytical Sequence
                                                                                      *
              All  samples and required blanks must be analyzed within a valid
              analysis sequence as given below.
                         Iniection
                                                  Material Injected
      0 hr.
       12 hr.
1-15

16

17

18
0
0
0
0
First 15 steps of the initial
calibration
Instrument blank at end of
initial calibration
PEM at end of initial
calibration
First sample


Subsequent samples


Last sample
                                         D-39/PEST
                                                          OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure (Con't)
      Time
      Another 12 hr.
      Another 12 hr.
In-jection #

1st injection past
12:00 hr.
2nd and 3rd
injections past
12:00 hr.
0
0
0
0
0
0
1st injection past
12 hr.
2nd injection
0
0
0
0
0
0
1st injection past
12:00 hr.
2nd and 3rd
injections past
12:00 hr.
0
0
0
0
0
etc.
Material Injected

Instrument blank

Individual Standard Mixtures
A and B

Sample

Subsequent samples
Last sample
Instrument blank

PEM
Sample

Subsequent samples
Last sample
Instrument blank

Individual Standard Mixtures
A and B

Sample

Subsequent samples
10.2.2.1.1       The first 12 hours are counted from injection #16  (the
                 Instrument Blank at the end of the initial calibration
                 sequence), not  from injection #1.  Samples and required blanks
                 may be injected until 12:00 hours have elapsed.  All subsequent
                 12-hour periods are timed from the injection of the instrument
                 blank that brackets the front end of the samples.  Because the
                 12-hour time period is timed from injection of the instrument
                 blank until the injection of the last sample, each 12-hour
                 period may be separated by the length of one chromatographic
                 run, that of the analysis of the last sample.  While the 12-
                 hour period may not be exceeded, the laboratory may run
                 instrument blanks and standards more frequently, for instance,
                 to accommodate  staff working on 8-hour shifts.
10.2.2.1.2       After the initial calibration, the analysis sequence may
                 continue as long as acceptable instrument blanks, PEMs and
                 Individual Standard Mixtures A and B are analyzed at the
                 required frequency.  This analysis sequence shows only the
                 minimum required blanks and standards.  More blanks and
                 standards may be run at the discretion of the Contractor;
                 however, the blanks and standards must also satisfy the
OLC03.2
                D-40/PEST

-------
                                              Exhibit D Pesticides/Aroclors -- Section 10
                                                                        Procedure  (Con't)

                 criteria presented in Section 9 in order to continue the run
                 sequence.

10.2.2.1.3           An analysis sequence must  also include  all  samples  and
                     required blank analyses, but the  Contractor may decide  at
                     what point in the sequence they are to  be analyzed.

10.2.2.1.4           The requirements  for the analysis sequence  apply to each GC
                     column and for all instruments used for these analyses.

10.2.3    Sample Dilutions

10.2.3.1      The sample must first be analyzed at the most  concentrated level
              (injection taken from the 2.0 mL  final extract after the clean-up
              steps).

10.2.3.2      If the concentration of any single component pesticide is greater
              than the upper limit of the initial calibration range on both GC
              columns, then the extract must be diluted.  If the concentration
              of any single component pesticide is greater than the upper limit
              of the initial calibration range  on one GC column, but not the
              other, then the extract must be diluted only if the percent
              difference between the two concentrations is less than or equal to
              25%.  The on-column concentration of the pesticide compound(s) in
              the diluted extract must be between the initial calibration low
              point and high point standards.

10.2.3.3      If the calculated concentration of any multicomponent peak, used
              for quantitation is greater than the concentration of the most
              intense  single component analyte in the initial calibration high
              point  standard, then the sample must be diluted to have the
              concentration of the largest peak in the multicomponent analyte
              between  the initial calibration midpoint and high point standards
              of that  single component pesticide.

10.2.3.4      If dilution is employed solely to bring a peak within the
              calibration range or to get a multicomponent pattern on scale, the
              results  for both the more  and the less concentrated extracts  must.
              be reported.  The resulting changes  in quantitation limits and
              surrogate  recovery must be reported also for the diluted samples.

10.2.3.5      If the dilution  factor  is  greater than 10, an additional extract
              10 times more concentrated than  the  diluted sample extract must be
              analyzed and reported with the sample data.   If the dilution
              factor is  less than  10, but greater  than  1, the results of the
              original undiluted analysis must also be reported.

10.2.3.6      If the analysis  of the most concentrated extract does not meet  the
              requirement for  dilution in Section  10.2.3.2  and  10.2.3.3, then
              the analysis is  at no additional cost to USEPA.

10.2.3.7      When  diluted,  the chromatographic  data  for the single component
              pesticide  must be able  to  be reported at greater  than 10 percent
              of  full  scale  but less  than 100  percent of full scale.

10.2.3.8      When  diluted,  multicomponent analytes must be able  to be  reported
              at  greater than  25 percent of  full  scale  but  less  than  100 percent
              of  full  scale.

10.2.3.9      If  a  chromatogram is replotted electronically to  meet these
              requirements,  the scaling  factor used must be displayed on the
              chromatogram.   If the chromatogram of any sample  needs  to be
              replotted electronically to meet these  requirements,  both the

                                         D-41/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure (Con't)

              initial chromatogram and the replotted chromatogram(s)  must be
              submitted in the data package.

10.2.3.10     Dilute the sample using the following procedure:

10.2.3.10.1      Calculate the extract dilution in order for the single
                 component pesticides to meet the requirement listed in Section
                 10.2.3.7.

10.2.3.10.2      Calculate the extract dilution in order for the multicomponent
                 analytes to meet  the requirement listed in Section 10.2.3.8.

10.2.3.10.3      Dilute the sample extract quantitatively with hexane.
OLC03.2                                 D-42/PEST

-------
                                                       Exhibit D Pesticides -- Section 11
                                                           Data Analysis and Calculations

11.0   DATA ANALYSIS AND CALCULATIONS

11.1   Qualitative Identification of Target Compounds

11.1.1    The laboratory will identify single component analyte peaks based on
          the Retention Time  (RT) windows  established during the initial
          calibration sequence.  Single component analytes are identified when
          peaks are  observed in  the  RT window for the analyte on both Gas
          Chromatograph  (GC) columns.

11.1.2    A set of three to five major peaks is  selected for each
          multicomponent analyte.  RT windows for each peak are determined from
          the initial calibration analysis.  Identification of a multicomponent
          analyte in the sample  is based on pattern  recognition in conjunction
          with the elution of three  to five sample peaks within the RT window
          of the corresponding peaks of the standard on both GC columns.  The
          number.of  potential quantitation peaks is  listed in Table D-2.

11.1.3    A standard of any identified multicomponent analyte must be run
          within 72  hours of  its detection in a  sample chromatogram within a
          valid 12-hour sequence.

11.1.4    The choice of the peaks used for multicomponent analyte
          identification and  the recognition of  those peaks may be complicated
          by the environmental alteration  of the toxaphene or Aroclors,  and by
          the presence of coeluting  analytes, or matrix interferences, or both.
          Because of the alteration  of these materials in the environment,
          multicomponent analytes in samples may give patterns similar to, but
          not identical with, those  of the standards.

11.1.5    Toxaphene  and Aroclors require only a  single-point calibration.
           Identification requires visual inspection of an on-scale pattern.

11.2   Calculations

11.2.1    Target  Compounds

11.2.1.1      Quantitation for all analytes and  surrogates must be performed and
              reported for each  GC column.

11.2.1.2      Manual integration of  peaks  (e.g., measuring peak height with a
              ruler) is only permitted when accurate electronic integration of
              peaks  cannot be done.   If manual integration of peaks is required,
              it must be documented  in the Sample Delivery Group (SDG)
              Narrative.

11.2.1.3      The Contractor must quantitate each single component analyte and
              surrogate based on the calibration factor from the most recent
              initial calibration midpoint standard mixture analyses.  Do not
              use the analyses of the Individual Standard Mixtures used to
              demonstrate calibration verification for quantitation of samples.

11.2.1.4      The Contractor must quantitate each multicomponent analyte based
              on the calibration factor from the most recent initial calibration
              standard.

11.2.1.5      If more than one multicomponent analyte is present, the Contractor
              must choose separate peaks to quantitate the different
              multicomponent analytes.  A  peak common to both analytes present
              in the sample must not be used to  quantitate either analyte.
                                         D-43/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 11
Data Analysis and Calculations  (Con't)

11.2.1.6      Before reporting data to USEPA, it is required that the Contractor
              check for flags generated by the data system that indicate
              improper quantitation of analytes.

11.2.1.7      The chromatograms of all samples  (including Laboratory Control
              Samples, requested matrix spike/matrix spike duplicate) standards,
              and required blanks must be reviewed by a qualified pesticide
              analyst before they are reported.

11.2.1.8      Calculate the concentration of the single component pesticides and
              surrogates by using the following equation:

              EQ. 12

                                                   |1J (Vfc) (Bt)
                          Concentration
                                                  CFtap
              Where:

               AX =    Response (peak area or height)  of the compound to be
                       measured.

               CFmp=    Calibration factor for the mid-point from initial
                       calibration standard (area per ng).

     •••"'         Vt =    Volume of concentrated extract (uL) .   (This volume is
                       2000 uL).

     •'.'••        Vi =    Volume of extract injected (uL) .   (If a single injection
                       is made onto two columns, use one-half the volume in the
     >:F>-                syringe as the volume injected onto each column) .
     »?.?.
               Vx =    Volume of water extracted (mL).  (NOTE:  for instrument
                       blanks and sulfur cleanup blanks, assume a 1,000 mL
                       volume).

               Df =    Dilution factor.  The dilution factor for analysis of
                       water samples by this method is defined as follows:

      uL most concentrated  extract used  to make dilution + uL clean  solvent
                    uL most concentrated extract used to make dilution

              If no dilution is performed, Df =  1.0.

              The calibration  factors used in Equation 12 are those from the
              most recent mid-point standard from the initial calibration.  If
              the calibration  factors used to determine the linearity of the
              initial calibration were based on peak area,  then the
              concentration of  the analyte in the sample must be based on peak
              area.  Similarly, if peak height was used to determine linearity,
              use peak height  to determine the concentration in the sample.

11.2.1.9      Contract Required Quantitation Limit  (CRQL)  Calculation

              If the adjusted  CRQL is less than  the CRQL listed in Exhibit C
              (Pesticides), report the CRQL in Exhibit C (Pesticides).
              Calculate the adjusted CRQL for pesticides by using the following
              equation:
OLC03.2                        '          D-44/PEST

-------
                                                      Exhibit D Pesticides -- Section 11
                                                  Data Analysis and Calculations  (Con't)
             EQ. 13
                                    Contract Stomp le
            Adjusted   Contract    Vol (1000 nL>     M          Vt
              CRfiL    "   CROL    *        v                 Contract Ext.
                                                              Vol  (2000uL)
             Where :

               Contract  CRQL  =     Take exact CRQL values reported in Exhibit C
                                  of the SOW.

               Vx  =               Same as EQ. 12.

               Vc  =               Same as EQ. 12.

               Df   =               Same as EQ . 12 .

11.2.1.10    During  initial  calibration, a set of three to five quantitation
             peaks was  chosen  for each multicomponent analyte.  Calculate the
             concentration of  each of the selected Aroclor or toxaphene peaks
             individually using  Equation 12.  Determine the mean concentration
             for  all of the  selected peaks.  The mean value is reported on Form
             X (Exhibit B) for both GC columns.

11.2.1.11    For  the single  component pesticides, report the lower of the two
             values  quantitated  from the two GC  columns of Form I .  For the
             multicomponent  analytes, report the lower of the two mean values
             from the two GC columns on Form I .

11.2.1.12    The  percent difference is calculated according to Equation 14.

             EQ.  14

                                     Cone. - Cone.
                                  • - — - — * 100
              Where :

               ConcH  =    The higher of the two concentrations for the target
                          compound in question.

               ConcL  =    The lower of the two concentrations for the target
                          compound in question.

              NOTE: Using this equation will result  in percent difference values
              that are always positive.   The value will also  be greater than a
              value calculated using the higher concentration in  the
              denominator,  however,  given the likelihood of a positive
              interference raising the concentration determined on one GC
              column, this is a conservative approach to comparing the two
              concentrations .

11.2.2     Surrogate  Recoveries

11.2.2.1      The concentrations of the surrogates are calculated separately for
              each GC column in a similar manner as  the other analytes using
              Equation 12 .   Use the calibration factors from  the  midpoint
              concentration of Individual Standard Mixture A  from the initial
              calibration.
                                        D-45/PEST                                 OLC03.2

-------
Exhibit D Pesticides /Aroclors -- Section 11
Data Analysis and Calculations  (Con't)

11.2.2.2      The recoveries of the surrogates are calculated according to
              Equation 15.
              EQ.  15
                              Percent Recovery • — - * 100
              Where:

               Qd =        Quantity determined by analysis.

               Qa =        Quantity added to sample/blank.

11 . 3   Technical Acceptance Criteria for Sample Analyses

       The requirements below apply independently to each GC column and to all
       instruments used for these analyses.  (See exception in Section 11.3.7)
       Quantitation must be performed on each GC column.

11.3.1    Samples  must be  analyzed under the Gas Chromatograph/Electron Capture
          Detector (GC/ECD) operating conditions in Section 9.1.  The
          instrument  must  have met all  initial  calibration and calibration
          verification technical  acceptance criteria.   Samples must be cleaned-
          up using Florisil that  meets  the technical acceptance criteria for
          Florisil.   Sample data  must be bracketed at  12-hour intervals  (or
          less)  by acceptable analyses  of  instrument blanks. Performance
          Evaluation  Mixtures  (PEMs) , and  Individual Standard Mixtures A and B,
      '    as described in  Section 10.2.2.1.

11.3.2'r   The  sample  must  be extracted  and analyzed within the contract holding
          times .

11.3.3    The  Laboratory Control  Sample (LCS) associated  with the samples must
     .;:•.   meet the LCS technical  acceptance criteria.

11.3.4    The  samples must have an associated method blank meeting the method
          blank  technical  acceptance criteria.  If a sulfur cleanup blank is
          associated  with  the samples,  that blank must meet the sulfur cleanup
          blank  technical  acceptance criteria.

11.3.5    The  RT for  each  of the  surrogates must be within the RT window as
          calculated  in Section 9.2.4.3, for both GC columns.

11.3.6    The  percent recovery for the  surrogates must be between 30.0 and 150
          percent,  inclusive.  These limits are not advisory.

          NOTE:  The surrogate recovery  requirements do not apply to a sample
          that has been diluted.

11.3.7    No target analyte concentrations may  exceed  the upper limit of the
          initial  calibration or  else extracts  must be diluted and re-analyzed.
          If a target analyte concentration exceeds the upper limit of the
          initial  calibration on  one GC column, but not the other, the extract
          must be  diluted  and re-analyzed only  if the  percent difference
          between  the two  concentrations is less than  or  equal to 25%.

11.3.8    A standard  for any identified multicomponent analyte must be analyzed
          on the same instrument  within 72 hours of its detection in a sample
          within a valid 12 hour  sequence.
OLC03.2                                 D-46/PEST

-------
                                                       Exhibit D Pesticides -- Section 11
                                                   Data Analysis and Calculations (Con't)

11.3.9    The identification of single component pesticides by gas
          chromatographic methods is based primarily on RT data.  The RT of the
          apex of a peak can only be verified from an on-scale chromatogram.
          The identification of multicomponent analytes by gas chromatographic
          methods is based primarily on recognition of patterns of RTs
          displayed on a chromatogram.  Therefore, the following requirements
          apply to all data presented for single component and multicomponent
          analytes.

11.3.9.1      When no analytes are identified in a sample,  the chromatograms
              from the analyses of the sample extract  must  use the  same scaling
              factor as was used for the low point standard of the  initial
              calibration associated with those analyses.

11.3.9.2      Chromatograms must display single component pesticides detected in
              the sample at less than full scale.

11.3.9.3      Chromatograms must display the largest peak of any multicomponent
              analyte detected in the sample at less than full scale.

11.3.9.4      If an extract must be diluted, chromatograms  must display single
              component pesticides between 10 and 100 percent of full scale.

11.3.9.5      If an extract must be diluted, chromatograms  must display
              multicomponent analytes between 25 and 100 percent of full scale.

11.3.9.6      For any sample or blank, the baseline of the chromatogram must
              return to below 50 percent of full scale before the elution time
              of alpha-BHC, and return to below 25 percent of full scale after
              the elution time of alpha-BHC and before the elution time of
              decachlorobiphenyl.

11.3.9.7      If a chromatogram is replotted electronically to  meet these
              requirements, the scaling factor-used must be displayed on the
              chromatogram.

11.3.9.8      If the chromatogram of any sample needs to be replotted
              electronically to meet these requirements, both the initial
              chromatogram and the replotted chromatogram(s) must be submitted
              in the data package.

11.4   Corrective Action for Sample Analysis

11.4.1     Sample analysis technical acceptance criteria MUST be met before data
           are reported.   Samples contaminated from laboratory sources  or
           associated with a contaminated method blank or  sulfur cleanup blank
           will require re-extraction and re-analysis  at no additional  cost to
           USEPA.   Any samples analyzed that do not meet the technical
           acceptance criteria will require re-extraction and/or re-analysis at
           no additional cost to USEPA.

11.4.2     If the sample analysis technical acceptance criteria are not met,
           check calculations,  surrogate solutions,  and instrument  performance.
           It may be necessary to recalibrate the instrument or take other
           corrective action procedures to meet the technical acceptance
           criteria,  in which case,  the affected samples must be re-analyzed .at
           no additional cost to USEPA after the corrective action.

11.4.3.    If the Contractor needs to analyze more than the most concentrated
           extract and two (2)  sample dilutions to have all the
           pesticide/Aroclor compounds within the calibration range of  the
           instrument, contact Sample Management Office (SMO).  SMO will contact
           the Region for instructions.

                                         D-47/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 11
Data Analysis and Calculations  (Con't)

11.4.4    Sample re-extraction/re-analyses performed as  a  result  of  suspected
          matrix interferences  beyond the  scope of  the method  will be  reviewed
          on a  case-by-case  basis  for payment purposes by  the  USEPA  Contract
          Laboratory Program Project  Officer (CLP PO).   Send a copy  of the  SDG
          Narrative  (including  your contract number), a  description  of the
          situation,  and the requested action to the CLP PO.
OLC03.2                                 D-48/PEST

-------
                                              Exhibit D Pesticides/Aroclors -- Section 12
                                                                          Quality Control

12.0   QUALITY CONTROL

12.1   Blank Analyses

12.1.1    Introduction

          There are  two types of blanks required by this method: the method
          blank and  the instrument blank.  A'separate sulfur cleanup blank may
          also be required if some, but not all of the samples are subjected to
          sulfur cleanup.  Samples that are associated with a sulfur cleanup
          blank are  also associated with the method blank with which they were
          extracted.  Both the method and sulfur cleanup blanks must meet the
          respective technical acceptance criteria for the sample analysis
          technical  acceptance criteria to be met.

12.1.2    Method Blank

12.1.2.1      Summary of Method Blank

              A method blank is 1.0 liter of reagent  water carried through the
              entire analytical scheme.   The purpose  of a method blank is to
              determine the levels of contamination associated with the
              processing and analysis of samples.

12.1.2.2      Frequency of Method Blank

              A method blank must be extracted each time samples are extracted.
              The number of samples extracted with each method blank shall not
              exceed 20 field samples (excluding matrix spike/matrix spike
              duplicate, PE samples,  and Laboratory Control Samples).   In
              addition,  a method blank shall:

              •    Be extracted by the same  procedure  used to  extract  samples;
                   and

              •    Be analyzed on each Gas Chromatograph/Electron Capture
                   (GC/ECD)  system used  to analyze associated  samples.

12.1.2.3      Procedure for Method Blank

              Measure 1.0 liter of reagent water for each method blank aliquot.
              Add 200 uL of the surrogate solution (Section 7.2.3.1).   Extract,
              concentrate and analyze the method blank according to Section 10.

12.1.2.4      Calculations for Method Blank

              Calculate method blank results according to Section 11.

12.1.2.5      Technical Acceptance Criteria for Method Blank

12.1.2.5.1       The requirements below apply independently to  each GC column
                 and to all  instruments used for  these analyses.   Quantitation
                 must be performed on each  GC column.

12.1.2.5.2       All method blanks must be  prepared  and analyzed at the
                  frequency described in Section 12.1.2.2,  using the procedure in
                  Section 12.1.2.3 on a  GC/ECD system meeting  the initial
                  calibration and calibration verification technical acceptance
                  criteria.   Method blanks must be cleaned-up  using Florisil
                 meeting the technical  acceptance criteria for  florisil.

12.1.2.5.3       Method blanks must be  bracketed at  12-hour  intervals (or less)
                 by acceptable analyses of  instrument blanks,  PEMs, and

                                         D-49/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 12
Quality Control  (Con't)

                 individual Standard Mixtures A and B as described  in  Section
                 10.2.2.1.

12.1.2.5.4       The concentration of  the  target  compounds  (Exhibit C  -
                 Pesticides)in  the method  blank must be less  than the  Contract
                 Required Quantitation Limit  (CRQL) for each  target compound.

12.1.2.5.5       The method blank must meet all sample technical acceptance
                 criteria in Sections  11.3.5 to 11.3.9.

12.1.2.5.6       Surrogate recoveries  must fall within the  acceptance  window of
                 30-150 percent.  These limits  are not advisory.

12.1.2.6      Corrective Action for Method Blank

12.1.2.6.1       If a  method blank does not meet  the technical acceptance
                 criteria the Contractor must consider the  analytical  system to
                 be out of control.

12.1.2.6.2       If contamination is a problem, the source  of the contamination
                 must  be investigated  and  appropriate corrective measures  MUST
                 be taken and documented before further sample analysis
                 proceeds.  Further, all samples  (including Laboratory Control
                 Samples, requested matrix spike/matrix spike duplicate, and PE
                 samples) processed with a method blank that  does not  meet the
                 method blank technical acceptance criteria (i.e.,  contaminated)
       -          will  require re-extraction and re-analysis at no additional
       ,f          cost  to USEPA.  It is the Contractor's responsibility to  ensure
      .»V-         that  method interferences caused by contaminants in solvents,
      •-•£          reagents, glassware,  and  sample  storage and  processing  hardware
                 that  lead to discrete artifacts  and/or elevated baselines in
                 gas chromatograms be  eliminated.

12.1.2.6.3       If surrogate recoveries in the method blank  do not meet the
                 technical acceptance  criteria, listed in 12.1.2.5.6,  first  re-
      "&         analyze the method blank.  If  the surrogate  recoveries  do not
      '•*         meet  the technical acceptance  criteria after re-analysis, then
                 the method blank and  all  samples (including  Laboratory  Control
                 Samples, requested matrix spike/matrix spike duplicate, and PE
                 samples) associated with  that  method blank must be re-extracted
                 and re-analyzed at no additional cost to USEPA.

12.1.2.6.4       If the method  blank fails to meet a technical acceptance
                 criteria other than Sections 12.1.2.5.4 and  12.1.2.5.6, then
                 the problem is an instrument problem.  Correct the instrument
                 problem, recalibrate  the  instrument  (if necessary) and  re-
                 analyze the method blank.

12.1.3     Sulfur Cleanup Blank

12.1.3.1      Summary of Sulfur Cleanup Blank

              The sulfur cleanup blank  is a modified form of the method blank.
              The sulfur cleanup blank  is hexane spiked with the surrogates and
              passed through the sulfur cleanup and analysis procedures.   The
              purpose of the sulfur cleanup is to determine the levels of
              contamination associated with the separate sulfur cleanup-steps.

12.1.3.2      Frequency of Sulfur Cleanup Blank

              The sulfur cleanup blank  is prepared when only part of a set of
              samples extracted together requires sulfur removal.   A method
              blank is associated with  the entire set of samples.   The sulfur

OLC03.2                                  D-50/PEST

-------
                                              Exhibit D Pesticides/Aroclors -- Section 12
                                                                  Quality Control (Con't)

              cleanup  blank is  associated with the part  of  the  set  which
              required sulfur cleanup.   If all the samples  associated with a
              given method blank are subjected to sulfur cleanup, then no
              separate sulfur cleanup blank is required.

12.1.3.3      Procedure for Sulfur Cleanup Blank

              The concentrated  volume of the blank must  be  the  same as the final
              volume of the samples associated with  the  blank.   The sulfur blank
              must also contain the surrogates at the same  concentrations as the
              sample extracts (assuming 100.0 percent recovery).  Therefore, add
              0.2 milliliters (mL)  of the surrogate  spiking solution (Section
              7.2.3.1) to 1.8 mL of h'exane in a clean vial.

12.1.3.3.2       Proceed with the  sulfur  removal (Section  10.1.5.3). using the
                 same  technique  (mercury  or  copper)  as the samples associated
                 with the  blank.

12.1.3.3.3       Analyze the sulfur blank according to Section 10.2.

12.1.3.4      Calculations for Sulfur Cleanup Blank

12.1.3.4.1       Assuming  that  the material  in  the sulfur  blank resulted from
                 the  extraction of a 1  L  water  sample, calculate the
                 concentration  of  each  analyte  using Equation 12 in  Section
                 11.2.1.8.   Compare  the results to the CRQL values in  Exhibit C
                  (Pesticides).

12.1.3.4.2       See  Section 11.2  for the equations  for  the other  calculations.

12.1.3.5    - Technical Acceptance•Criteria for Sulfur Cleanup.Blanks

12.1.3.5.1     '  The-requirements  below apply independently to each  GC column
                 and  to all instruments used for.these  analyses.   Quantitation
                 must be performed on each  column.                        -

12.1.3.5.2       All  sulfur cleanup blanks'must be prepared and analyzed at the
                  frequency described in Section 12.1.3.2 using the procedure in
                  Section 12.1.3.3  on a  GC/ECD system meeting the initial
                  calibration and calibration verification  technical  acceptance
                  criteria.

12.1.3.5.3        Sulfur cleanup blanks  must be bracketed at 12-hour  intervals
                  (or  less) by  acceptable analyses of  instrument blanks,  PEMs and
                  Individual Standard Mixtures A and  B,  as  described  in Section
                  10.2.2.1.

12.1.3.5.4        The  concentration of  the target compounds (Exhibit  C -
                  Pesticides) in the sulfur cleanup blank must be less than the
                  CRQL for each target  compound.

12.1.3.5.5        The  sulfur cleanup blank must meet  all sample  technical
                  acceptance criteria in Sections 11.3.5 to 11.3.9.

12.1.3.5.6        Surrogate recoveries  must fall within  the acceptance windows of
                  30-150 percent.

12.1.3.6      Corrective Action for  Sulfur  Cleanup Blank

12.1.3.6.1        If a sulfur cleanup blank does not  meet the  technical
                  acceptance criteria,  the Contractor must  consider the
                  analytical system to  be out of control.


                                         D-51/PEST                                  OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 12
Quality Control (Con't)

12.1.3.6.2       If contamination is a problem, the source of the contamination
                 must be investigated and appropriate corrective measures MUST
                 be taken and documented before further sample analysis
                 proceeds.  Further, all samples processed with a sulfur cleanup
                 blank that does not meet the sulfur cleanup blank technical
                 acceptance criteria  (i.e., contaminated) will require re-
                 extraction and re-analysis at no additional cost to USEPA.  It
                 is the Contractor's responsibility to ensure that method
                 interferences caused by contaminants in solvents, reagents,
                 glassware, and sample storage and processing hardware that lead
                 to discrete artifacts and/or elevated baselines in gas
                 chromatograms be eliminated.

12.1.3.6.3       If surrogate recoveries in the sulfur cleanup blank do not meet
                 the technical acceptance criteria, in Section 12.1.3.5.6, first
                 re-analyze the sulfur cleanup blank.  If the surrogate
                 recoveries do not meet the technical acceptance criteria after
                 re-analysis, then the sulfur cleanup blank and all samples
                 associated with that sulfur cleanup blank must be re-
                 prepared/re-extracted and re-analyzed at no additional cost to
                 USEPA.

12.1.3.6.4       If the sulfur cleanup blank fails to meet a technical
                 acceptance criteria other than Sections 12.1.3.5.4, and
                 12.1.3.5.6, then the problem is an instrument problem.  Correct
                 the instrument problem, recalibrate the instrument (if
                 necessary) and re-analyze the sulfur cleanup blank.

12.1.4'    Instrument Blank

12.1..4-1      Summary of Instrument Blank

     • '•&,       An instrument blank is a volume of clean solvent spiked  with the
     -iX-       surrogates and analyzed on each GC column and instrument used for
     •-•        sample analysis.   The purpose of the instrument blank is to
              determine the levels of contamination associated with the
     •%       instrumental analysis particularly with regard to carryover of
              analytes from standards or highly contaminated samples into other
              analysis.

12.1.4.2      Frequency of Instrument Blank

              The first analysis in a 12-hour analysis sequence (Section 9.3.2)
              must be an instrument blank.   All groups of acceptable sample
              analyses are to be preceded and followed by acceptable instrument
              blanks (Section 10.2.2.1).   If more than 12 hours have elapsed
              since the injection of the instrument blank that bracketed a
              previous 12-hour period,  an instrument blank must be analyzed to
              initiate a new 12-hour sequence (Section 9.3.2).

12.1.4.3      Procedure for Instrument Blank

12.1.4.3.1       Prepare the instrument blank by spiking the surrogates into
                 hexane or iso-octane for a concentration of 20.0 nanograms per
                 milliliter  (ng/mL) of Tetrachloro-m-xylene and
                 Decachlorobiphenyl.

12.1.4.3.2       Analyze the instrument blank according to Section 10.2 at the
                 frequency listed in Section 12.1.4.2.
OLC03.2                                 D-52/PEST

-------
                                              Exhibit D Pesticides/Aroclors -- Section 12
                                                                  Quality Control (Con't).

12.1.4.4      Calculations for Instrument  Blank

12.1.4.4.1       Assuming that the material in the instrument blank resulted
                 from the extraction of a 1 L water sample, calculate the
                 concentration of each analyte using Equation 12 in Section
                 11.2.1.8.  Compare the results to the CRQL values for water
                 samples in Exhibit C  (Pesticides).

12.1.4.4.2       See Section  11.2 for the equations for the other calculations.

12.1.4.5      Technical  Acceptance Criteria for Instrument Blanks

12.1.4.5.1       The requirements'below apply independently to each GC column
                 and to all instruments used for these analyses.  Quantitation
                 must be performed on each GC column.

12.1.4.5.2       All instrument blanks must be prepared and analyzed at the
                 frequency described in Section 12.1.4.2  using the procedure in
                 Section  10.2 on  a GC/ECD system meeting  the initial calibration
                 and calibration  verification technical acceptance criteria.

12.1.4.5.3       The concentration of  each target analyte (Exhibit C -
                 Pesticides)  in the instrument blank must be less than the CRQL
                 for that analyte.

12.1.4.5.4       The instrument blank  must meet all sample technical acceptance
                 criteria in  Sections  11.3.5 to 11.3.9.

12.1.4.6      Corrective Action for Instrument Blank

12.1.4.6.1       If analytes  are  detected at concentrations greater than  the
                 CRQL or  the  surrogate Retention  Times  (RTs) are outside  the RT
                • windows,  all data collection must-be stopped, and corrective
                 action must  be taken.  Data for. samples.which were run between
                 the last-acceptable instrument blank and the unacceptable blank
                 are considered suspect.  An acceptable instrument blank  must be
                 run before additional data are collected. All  samples
                  (including Laboratory Control Samples, requested matrix
                 spike/matrix spike duplicate, and  PE samples) and required
                 blanks which were run after the  last acceptable instrument
                 blank  must be reinjected during  a  valid  run sequence  and must
                 be reported  at no additional  cost  to USEPA.

12.2    Laboratory Control Sample  (LCS)

12.2.1     Summary  of  LCS

           The LCS  is  an internal  laboratory quality control  sample  designed to
           assess [on  a Sample Delivery Group (SDG)-by-SDG basis] the  capability
           of the contractor to perform the analytical  method listed in this
           Exhibit.

12.2.2     Frequency of LCS

           The LCS  must be prepared,  extracted,  analyzed,  and reported once per
           Sample Delivery Group.   The LCS must be extracted and analyzed
           concurrently with the samples  in the SDG using the same  extraction
           protocol and instrumentation as the samples  in the SDG.
                                         D-53/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors  -- Section 12
Quality Control  (Con't)

12.2.3    Procedure for LCS    >

          Measure a 1  liter  aliquot  of reagent water in a 1 liter graduated
          cylinder and transfer  the  water  to a continuous extractor or 2 L
          separatory funnel.   Pipet  1.0 mL of the LCS spiking solution (Section
          7.2.3.8}  and 200 uL of the surrogate standard spiking solution into
          the water and mix  well.  Extract,  concentrate,  and analyze the sample
          according to Section 10.

12.2.4    Calculations for LCS

12.2.4.1      Calculate the results according to Section  11.

12.2.4.2      Calculate individual compound recoveries of the LCS using Equation
              15.

12.2.5    Technical Acceptance Criteria For LCS

12.2.5.1      The requirements below apply independently  to each GC column and
              to all instruments used for these analyses.  Quantitation must be
              performed on each GC column.

12.2.5.2      The LCS must be analyzed at the frequency described in Section
              12.2.2 on a GC/ECD system meeting the initial calibration and
              calibration verification technical acceptance criteria.

12.2.5.3      The LCS must be prepared as described in Section  12.2.3.

12.2.5V4      The LCS must meet all sample technical acceptance criteria in
              Sections 11.3.5 to 11.3.9.
      Sfc."
12.2.5"S5      The percent recovery for each of the compounds in the LCS must be
      "Si--      within the recovery limits listed in Table  D-3.
      *xri*.
      ~>t~
12.2.6iT.>  Corrective Action  for  LCS
      -.**-•
12.2.6V1      If the LCS technical acceptance criteria for the  surrogates.or the
              LCS compound recovery are not met, check calculations, the
              surrogate and LCS solutions,  and instrument performance.  It may
              be necessary to recalibrate the instrument  or take other
              corrective action procedures to meet the surrogate and LCS
              recovery criteria.

12.2.6.2      LCS technical acceptance criteria MUST be met before data are
              reported.  LCS contamination from laboratory sources or any LCS
              analyzed not meeting the technical acceptance criteria will
              require re-extraction and re-analysis of the LCS at no additional
              cost  to USEPA.

12.2.6.3      All samples  (including matrix spike/matrix  spike  duplicate and PE
              samples)  and required blanks, prepared and  analyzed in an SDG with
              an LCS that does not meet the technical acceptance criteria, will
              also  require re-extraction and re-analysis  at no  additional cost
              to USEPA.

12.3   Matrix Spike/Matrix Spike Duplicate  (MS/MSD)

12.3.1    Summary of MS/MSD

           In order to  evaluate the effects of the sample matrix on the methods
          used for pesticide/Aroclor analyses,  USEPA has prescribed a mixture
          of pesticide/Aroclor target compounds to be spiked into two aliquots
          of a sample, and analyzed  in accordance with the appropriate method.

OLC03.2                                 D-54/PEST

-------
                                              Exhibit D Pesticides/Aroclors -- Section 12
                                                                  Quality Control (Con't)
12.3.2    Frequency of MS/MSD Analysis
12.3.2.1      MS/MSD samples shall only be analyzed if requested by the  Region
              [through Sample Management Office (SMO)3  or specified on the
              Traffic Report (TR).  If requested,  a matrix spike and matrix
              spike duplicate must be extracted and analyzed for every 20 field
              samples in an SDG.

12.3.2.2      As part of USEPA's  QA/QC program, water rinsate samples and/or
              field blanks may be delivered to a laboratory for analysis.   Do
              not perform MS/MSD  analysis on a water rinsate sample or field
              blank.

12.3.2.3      If the USEPA Region requesting MS/MSD designates a sample  to  be
              used as an MS/MSD,  then that sample must be used.  If there is
              insufficient sample volume remaining to perform an MS/MSD, then
              the Contractor shall choose.another sample to perform an MS/MSD"
              analysis.  At the time the selection is made, the Contractor  shall
              notify the Region (through SMO) that insufficient sample was
              received and identify the USEPA sample selected for the MS/MSD
              analysis.  The rationale for the choice of another sample  other
              than the one designated by USEPA shall be documented in the SDG
              Narrative.

12.3.2.4      If there is insufficient sample volume remaining in any of the
              samples in an SDG to perform the requested MS/MSD, the Contractor
              shall immediately contact SMO to inform them of the problem.   SMO
              will contact the Region for instructions.  The Region will either
              approve that no MS/MSD be performed, or require that a reduced
              sample aliquot be used for the MS/MSD analysis.  SMO will  notify
              the Contractor of the Region's decision.  The Contractor shall
              document the decision in the SDG Narrative.

12.3.2.5      If it appears that the Region has requested MS/MSD analysis  at a
              greater frequency than specified in Section 12.3.2.1, the
              Contractor shall contact SMO.  SMO will contact the Region to
              determine which samples should have MS/MSD performed on them.  SMO
              will notify the Contractor of the Region's decision.  The
              Contractor shall document the decision in the SDG Narrative.   If
              this procedure is not followed,  the Contractor will not be paid
              for MS/MSD analysis performed at a greater frequency than required
              by the contract.

12.3.2.6      When a Contractor receives only  Performance Evaluation  (PE)
              samples,  no MS/MSD  shall be performed within that SDG.

12.3.2.7      When a Contractor receives  a PE  sample as part of a larger SDG, a
              sample other  than the PE sample  must be chosen for the requested
              MS/MSD analysis when the Region  did not designate a sample to be
              used  for  this purpose.

12.3.3     Procedure for Preparing MS/MSD

12.3.3.1      Measure  out two additional  1 L aliquots of the sample chosen for
              spiking.  Adjust the pH of  the samples  (if required) and  fortify
              each with 1 mL of the matrix spiking solution  (Section 7.2.3.2) .
              Using a  syringe or  volumetric pipet, add 200 uL  of the surrogate
              spiking  solution (Section  7.2.3.1) to each sample.  Extract,
              concentrate,  cleanup, and  analyze the matrix spike and matrix
              spike duplicate according  to Section 10.0.

12.3.3.2      Matrix  spike  and matrix spike  duplicate samples  must be analyzed
              at  the  same concentration  as the most concentrated extract for

                                         D-55/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 12
Quality Control (Con't)

              which the original sample results will be reported.   Do not
              further dilute the MS/MSD samples to get either spiked or
              nonspiked analytes within calibration range.

12.3.4    Calculations  for  MS/MSD"

          The percent  recoveries and the relative percent difference between
          the recoveries of each of the compounds in  the matrix spike and
          matrix spike  duplicate samples will  be 'calculated  and reported by
          using the  following equations:

          EQ. 16
                                                   AtACB — *tl>
                        B*trix Spike Recovery •     SL    * 100
          Where:

            SSR  =

            SR =

            SA =
                 Spike sample  result.

                 Sample result.

                 Spike added.
          EQ.  17
      •••v
                             IHSR -
                                     £ (HSR «• KSDR)
                                                      * 100
Where:

 RPD =

 MSR =

 MSDR :
                           Relative percent difference.

                           Matrix spike recovery.

                           Matrix spike duplicate recovery.
12.3.5    Technical Acceptance  Criteria  for MS/MSD

12.3.5.1      The requirements below apply independently to each GC column and
              to all instruments used for these analyses.   Quantitation must be
              performed on each GC column.

12.3.5.2      If requested, all MS/MSD must be prepared and analyzed at the
              frequency described in Section 12.3.2,  using the procedure above
              and in Section 10 on a GC/ECD system meeting the initial
              calibration, calibration verification,  and blank technical
              acceptance criteria.  MS/MSD must be cleaned-up using florisil
              meeting the technical acceptance criteria for florisil.   MS/MSD
              must be bracketed at 12-hour intervals  (or less)  by acceptable
              analyses of instrument blanks,  PEMs, and individual standard
              mixtures A and B as described in Section 10.2.2.1.

12.3.5.3      The samples must be extracted and analyzed within the contract
              required holding times.

12.3.5.4      The RT for each of the surrogates must  be within the RT window as
              calculated in Section 9 for both GC columns.
OLC03.2
                              D-56/PEST

-------
                                              Exhibit D Pesticide's/Aroclors -- Section 12
                                                                  Quality Control (Con't)

12.3.5.5      The limits for matrix spike compound recovery and RPD are given in
              Table D-4.   As these limits are only advisory,  no further action
           .   by the laboratory is required.   However,  frequent failures to meet
              the limits for recovery or RPD warrant investigation by the
              laboratory,  and may result in questions from USEPA.

12.3.6    Corrective Action for MS/MSD

          Any MS/MSD which  fails  to meet  the technical acceptance  criteria  in
          Sections  12.3.5.1 through 12.3.5.4 must  be re-analyzed at  no
          additional cost to USEPA.

12.4   Method Detection Limit  (MDL) Determination

12.4.1    Before any field  samples are  analyzed under this contract,  the MDL
          for each pesticide target compound shall be determined for each
          sample extraction procedure and on one  of 'the  instruments  to be used
          for sample analysis.  The MDLs  must be  verified annually thereafter
           (see  Section  12.4.2 for MDL verification procedures), until the
          contract expires  or is  terminated or after major instrument
          maintenance.  Major instrument  maintenance includes,  but is not
          limited to replacement  of gas chromatographic  column or  replacement
          of  the electron capture detector.

12.4.2    To  determine  the  MDLs,  the  Contractor shall run an MDL  study
          following the procedures specified in 40 CFR Part 136.   The
          Contractor is only required to analyze  the'MDL samples  on  one
          instrument used for field sample analyses.  MDL verification only is
          then  required on  all other instruments  used  for field sample analysis
          and at the frequency^specified in Section 12.4.1.   MDL verification
          is  achieved by analyzing a;'single reagent,water blank spiked with
          each  target compound at-a concentration equal  to two times .the
          analytical determined MDL.  The resulting chromatogram.must meet  the
          qualitative identification criteria outlined in Sections 11.1.1
          through 11.1.5.

12.4.3    The determined concentration  of the MDL must be less than the CRQL.

12.4.4    All documentation for the MDL studies  shall  be maintained at the
           laboratory and provided to USEPA upon  written  request.
                                         D-57/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors - Sections 13-16
Method Performance

13.0   METHOD PERFORMANCE

       Not Applicable.

14.0   POLLUTION PREVENTION

14.1   Pollution prevention encompasses any technique that reduces or
       eliminates the quantity or toxicity of waste at the point of generation.
       Numerous opportunities for pollution prevention exist in laboratory
       operation.  The USEPA has established a preferred hierarchy of
       environmental management techniques that places pollution prevention as
       the management option of first choice.  Whenever feasible, laboratory
       personnel should use pollution prevention techniques to address their
       waste generation.  When wastes cannot be feasiblely reduced at the
       source, USEPA recommends recycling as the next best option.

14.2   For information about pollution prevention that may be applicable to
       laboratories and research institutions, consult Less is Better:
       Laboratory Chemical Management for Waste Reduction, available from the
       American Chemical Society's Department of Government Relations and
       Science Policy, 1155 16th Street N.W., Washington, B.C. 20036, (202)872-
       4477.

15.0   WASTE MANAGEMENT

       The Environmental Protection Agency requires that laboratory waste
       management practices be conducted consistent with all applicable rules
       and regulations.  USEPA urges laboratories to protect the air, water,
      Sv.and land by minimizing and controlling all releases from hoods and bench
      ^"operations, complying with the letter and spirit of any sewer discharge
      .^permits and regulations, and by complying with all solid and hazardous
      .ijwaste regulations, particularly the hazardous waste identification rules
 '     'Sand land disposal restrictions.  For further information on waste
      ^management consult The Waste Management Manual for Laboratory Personnel,
       available from the American Chemical Society at the address listed in
       Section 14.2.
      •="rv
16.0   REFERENCES

       Not Applicable.
OLC03.2                                 D-58/PEST

-------
                                              Exhibit D Pesticides/Aroclors -- Section 17
                                                               Tables/Diagrams/Flowcharts
17.0   TABLES/DIAGRAMS/FLOWCHARTS

                                        TABLE D-l
                   RETENTION TIME WINDOWS  FOR  SINGLE AND MULTICOMPONENT
                                 ANALYTES  AND  SURROGATES.
                                               Compound Identification
                Compound                          Window  (minutes)
                alpha-BHC                               ±0.05
                beta-BHC                                ±0.05
                gamma-BHC                               ±0.05
                delta-BHC                               ±0.05
                Heptachlor                              ±0.05
                Aldrin                                  ±0.05
                alpha-Chiordane                         ±0.07
                gamma-Chiordane                         ±0.07
                Heptachlor epoxide                      ±0.07
                Dieldrin                                ±0.07
                Endrin                                  ±0.07
                Endrin aldehyde                         ±0.07
                Endrin ketone                    .       ±0.07
                4,4'-ODD                                ±0.07
                4,4'-DDE                                ±0.07
                4,4'-DDT                                ±0.07
                Endosulfan I                            ±0.07
                Endosulfan II                           ±0.07
                Endosulfan sulfate                      ±0.07
                Methoxychlor                            ±0.07
                Aroclors                                ±0.07
                Toxaphene                               ±0.07
                Tetrachloro-m-xylene                    ±0.05
                Decachlorobiphenyl	±0.10	
                                         D-59/PEST                                 OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 17
Tables/Diagrams/Flcwcnarts  {Con't)
                                        TABLE D-2

                          NUMBER OF POTENTIAL  QUANTITATION PEAKS
         Multicomponent Analyte
 No. of Potential Quantitation Peaks
         Aroclor 1016/1260
         Aroclor 1221
         Aroclor 1232
         Aroclor 1242
         Aroclor 1248
         Aroclor 1254
         Toxaphene	
                 5/5
                  3
                  4
                  5
                  5
                  5
                  4
OLC03.2
D-60/PEST

-------
                                     Exhibit D Pesticides/Aroclors — Section  17
                                              Tables/Diagrams/Flowcharts  (Con't)
                               TABLE D-3

                LABORATORY CONTROL  SAMPLE RECOVERY LIMITS

 Compound	% Recovery
 gamma-BHC                                            50-120
 Heptachlor epoxide                                   50-150
 Dieldrin                                             30-130
 4,4'-DDE                                             50-150
 Endrin                                               50-120
 Endosulfan sulfate                                   50-120
 gamma - Chi or dane	30-130
NOTE: The recovery limits for any of the compounds in the LCS may be
expanded at any time during the period of performance if USEPA
determines that the limits are too restrictive.
                                D-61/PEST             .                    OLC03.2

-------
Exhibit D Pesticides/Aroclors -- Section 17
Tables/Diagrams/Flowcharts  (Con't)

                                        TABLE D-4

               MATRIX SPIKE RECOVERY AND RELATIVE PERCENT DIFFERENCE LIMITS


 Compound                               % Recovery                       RPD

 gamma-BHC  (Lindane)                      56-123                          15

 Heptachlor                              40-131                          20

 Aldrin                                   40-120                          22

 Dieldrin                                 52-126                 .         18

 Endrin                                   56-121                          21

 4,4'-DDT                                 38-127                          27
OLC03.2                                  D-62/PEST

-------
                          EXHIBIT E





QUALITY ASSURANCE/QUALITY CONTROL PROCEDURES AND REQUIREMENTS
                             E-l                                     OLC03.2

-------
                           THIS PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                    E-2

-------
       Exhibit  E  --  Quality Assurance/Quality Control  Procedures  and Requirements

                                    Table of Contents

                                                                                    Page


1.0    OVERVIEW	5
       1.1    Quality Assurance/Quality Control (QA/QC) Activities   	 5

2.0    INTRODUCTION  	 6
       2.1    Quality Assurance/Quality Control (QA/QC) Program Components  	 6

3.0    GENERAL QUALITY ASSURANCE/QUALITY CONTROL (QA/QC) 'PRACTICES	-7

4.0    SPECIFIC QUALITY ASSURANCE/QUALITY CONTROL (QA/QC) PROCEDURES  	 8
  ,    4.1    Purpose	8
       4.2    Laboratory Audit and Intercomparison Study Program  	 8
       4.3    Annual Verification of Method Detection Limits (MDLs)   	 8
       4.4    Quality Assurance/Quality Control Measurements	8

5.0    QUALITY ASSURANCE PLAN  (QAP)  	 9
       5.1    Introduction   	 9
       5.2    Required Elements of a Quality Assurance Plan  	  10
       5.3    Updating and Submitting the Quality Assurance Plan	11
       5.4    Corrective Actions   	  12

6.0    STANDARD OPERATING PROCEDURES  (SOPs)  	 	  13
      ' 6.1'    Introduction   	13
       6.2 .    Format	14
       6.3    Requirements   	14
       6:4,    Updating and Submitting SOPs	16
       6.5    Corrective Action  	  18

7..0   .CONTRACT COMPLIANCE SCREENING  (CCS)   	19
       7.1    Overview	19
       7.2    CCS Results	19
       7.3    CCS Trend Report	  19
       7.4    Corrective Actions   	  19

8.0    ANALYTICAL STANDARDS REQUIREMENTS   .  .  .  .'	20
       8.1    Overview	  20
       8.2    Preparation of Chemical Standards from the Neat High
              Purity Bulk Material   	20
       8.3    Purchase of Chemical Standards Already in Solution  	  21
       8.4    Requesting Standards from the USEPA Standards Repository   	  23
       8.5    Documentation of the Verification and Preparation of
              Chemical Standards	  23
       8.6    Corrective Actions   	  24

9.0    DATA PACKAGE AUDITS   	  25
       9.1    Overview	25
       9.2    Responding to the Data Package Audit Report	25
       9.3    Corrective Actions   	  25

10.0   REGIONAL DATA REVIEW	26
       10.1   Overview   	  26

11.0   PROFICIENCY TESTING   	  26
       11.1   Performance Evaluation  (PE) Samples	.-	26
       11.2   Quarterly Blind  (QB) Audits  	  27
       11.3   Corrective Actions   	  29
                                            E-3                                     OLC03.2

-------
        Exhibit  E —  Quality Assurance/Quality Control  Procedures  and Requirements

                              Table of Contents  (Continued)

Section                                                                             Page

12.0   ON-SITE LABORATORY EVALUATIONS 	   30
       12.1   Overview   	30
       12.2   Quality Assurance On-Site Evaluation  	   30
       12.3   Evidentiary Audit	'	   30
       12.4   Discussion of the On-Site Team's Findings .  . .	   31
       12.5   Corrective Action Reports for Follow-Through to  Quality
              Assurance and Evidentiary Audit Reports  	   31
       12.6   Corrective Actions   	   32

13.0   ELECTRONIC DATA AUDITS	   33
       13.1   Overview   	   33
       13.2   Submission of the Instrument Electronic Data  	   35
       13.3   Responding to the Electronic Data Audit Report  	   35
       13.4   Corrective Actions   	   35
       13.5   Maintenance of the Magnetic Tape Storage Device  	   36

14.0   DATA MANAGEMENT	   37
       14.1   Overview	37
       14.2   Documenting Data Changes   	37
       14.3   Lifecycle Management Procedures  	   37
       14.4   Personnel Responsibilities   	   38
 OLC03.2                                    E-4

-------
                                                                   Exhibit E -- Section 1
                                                                                 Overview
1.0    OVERVIEW
       Quality Assurance (QA)  and Quality Control (QC)  are integral parts of •
       the U.S. Environmental  Protection Agency's (USEPA's)  Contract Laboratory
       Program (CLP).   The QA process consists of management review and
       oversight at the planning, implementation, and completion stages of the
       environmental data collection activity, and ensures that data provided
       are of the quality required.   The QC process includes those activities
       required during data collection to produce the data quality desired and
       to document the quality of the collected data.

1.1    Quality Assurance/Quality Control (QA/QC) Activities

       During the planning of  an environmental data collection program, QA
       activities focus on defining data quality criteria and designing a QC
       system to measure the quality of data being generated.  During the
       implementation of the data collection effort, QA activities ensure that
       the QC system is functioning effectively, and that the deficiencies
       uncovered by the QC system are corrected.  After environmental data are
       collected, QA activities focus on assessing the quality of data obtained
       to determine its suitability to support enforcement or remedial
       decisions.

1.1.1     This exhibit describes the overall  QA/QC operations and the processes
          by which the CLP meets the QA/QC objectives defined above.  This
          contract requires a variety of QA/QC  activities.  These contract
          requirements are the minimum QC operations necessary to satisfy the
          analytical requirements associated  with the determination of the
          different method analytes.  These QC  operations are designed to
          facilitate laboratory  comparison by providing USEPA with comparable
          data from all Contractors.  These requirements do not release the
          analytical Contractor  from maintaining.their own QC checks on method
          and  instrument performance.
                                           E-5                                    OLC03.2

-------
Exhibit E -- Section 2
Introduction

2.0    INTRODUCTION

       Appropriate use of data generated under the large range of analytical
       conditions encountered in environmental analyses requires reliance on
       the Quality Control (QC) procedures and criteria incorporated into the
       methods.   The methods in this contract have been validated on samples
       typical of those received by the laboratories in the Contract Laboratory
       Program (CLP).   However, the validation of these methods does not
       guarantee that they perform equally well for all sample matrices
       encountered.  Inaccuracies can also result from causes other than
       unanticipated matrix effects, such as sampling artifacts, equipment
       malfunctions, and operator error.  Therefore, the QC component of each
       method is indispensable.

       The data acquired from QC procedures are used to estimate and evaluate
       the information content of analytical results and to determine the
       necessity for or the effect of corrective action procedures.  The
       parameters used to estimate information content include precision,
       accuracy, and other quantitative and qualitative indicators.  In
       addition, QC procedures give an overview of the activities required in
       an integrated program to generate data of known and documented quality
       required to meet defined objectives.

 2.1    Quality Assurance/Quality Control  (QA/QC) Program Components

 2.1.1      The necessary components of a complete QA/QC program include internal
           QC criteria that demonstrate acceptable levels  of  performance,  as
           determined by QA review.   External review of-data  and procedures is
           accomplished by the monitoring activities of the National Program
           Office,  Regional Data Users,  Sample Management  Office (SMO),  and the
           Quality Assurance Technical Support (QATS)  Laboratory.  Each external
           review accomplishes a different-purpose.   These reviews are described
          "in specific sections of this exhibit.   Laboratory  evaluation samples,
           electronic data audits,  and data packages provide  an external QA
           reference for the program.   A Contractor on-site evaluation system is
           also  part of the external QA monitoring.   A feedback loop provides
           the results of the various review functions to the.Contractors
           through direct communications with the USEPA Regional CLP Project
           Officer (CLP PO).

 2.1.2      This  exhibit does not provide specific instructions for constructing
           QA Plans, QC systems, or a QA organization.  It is,  however,  an
           explanation of the QA/QC requirements of the program.  It outlines
           some  minimum standards for QA/QC programs.  It also includes specific
           items that are required in a Quality Assurance Plan  (QAP) and by the
           QA/QC documentation detailed in this contract.   Delivery of this
           documentation provides USEPA with a complete data  package which will
           stand alone, and limits the need for contact with the Contractor or
           with an analyst, at a later date, if some aspect of the analysis is
           questioned.

 2.1.3      In order to assure that the product delivered by the Contractor meets
           the requirements of the contract, and to improve interlaboratory data
           comparison, USEPA requires the following from the Contractor:

           •    Preparation of,  and adherence to,  a  written QAP,  the elements of
                which  are designated in Section 5;

          , •    Preparation of,  and adherence to,  Standard Operating Procedures
                (SOPs)  as described in Section 6;
 OLC03.2                                    E-6

-------
                                                              Exhibit E -- Sections 2 & 3
                                                                  General QA/QC Practices

          •    Adherence to the analytical methods and associated QC
               requirements specified in the contract;

          •    Verification of analytical standards and documentation of the
               purity of neat materials and the purity and accuracy of solutions
               obtained from private chemical supply houses;

          •    Submission of all raw data and pertinent documentation for
               Regional review;

          •    Participation in the analysis of laboratory evaluation samples,
               including. adherence to corrective action procedures;

          •    Submission, upon request, of instrument data tapes and applicable
               documentation for tape audits, including a copy of the Sample
               Data Package ;

          •    Participation in on- site laboratory evaluations,  including
               adherence to corrective action procedures; and

          •    Submission of all original documentation generated during sample
               analyses for USEPA review.

3.0    GENERAL QUALITY ASSURANCE/QUALITY CONTROL (QA/QC)  PRACTICES
           Contractor shall adhere to good laboratory practices for laboratory
       •cleanliness with regard to glassware and apparatus.  The Contractor
       "shall also adhere to good laboratory practices with regard to reagents,
       •solvents,  and gases.  For additional guidelines regarding these general
      .^laboratory procedures, see the Handbook for Analytical Quality Control
      '•-'•in Water and Wastewater Laboratories USEPA-600/4-79-019, USEPA
       '•Environmental Monitoring Systems Laboratory, Cincinnati, Ohio,  September
       -1982.
                                           E-7                                    OLC03.2

-------
Exhibit E -- Section 4
Specific QA/QC Procedures

4.0    SPECIFIC QUALITY ASSURANCE/QUALITY CONTROL (QA/QC)  PROCEDURES •

       The QA/QC procedures defined herein shall be used by the Contractor when
       performing the methods specified in Exhibit D.   When additional  QA/QC
       procedures are specified in the methods in Exhibit D, the Contractor
       shall also follow these procedures.

       NOTE: The cost of performing all QA/QC procedures specified in this
       Statement of Work (SOW) is included in the price of performing the bid
       lot.

4.1    Purpose

4.1.1      The  purpose  of this document is  to  provide a  uniform set  of
           procedures  for the analysis of organic  samples,  documentation of
           methods  and their performance, and  verification of  the  sample data
           generated.   The  program will also assist  laboratory personnel in
           recalling and defending their actions under cross examination if
           required to present court  testimony in  enforcement  case litigation.
           Although it  is impossible  to address all  analytical situations in one
           document,  the approach taken here is to define  the  minimum
           requirements for all major steps relevant to  any organic  low
           concentration analysis.

4.1.2      The  primary function of the QA/QC program is  the definition  of
           procedures for the evaluation and documentation of  analytical
           methodologies and the reduction  and reporting of data.  The  objective
           is to provide a  uniform basis for sample  handling,  instrument and
           methods  maintenance, performance evaluation,  and analytical  data
           gathering and reporting.   In many instances where methodologies are
           available,  specific QC procedures are incorporated  into the  method
           documentation (Exhibit D).

4.2    Laboratory Audit and Intercomparison Study Program

       The  Contractor  is required to participate  in the Laboratory Audit  and
       Intercomparison Study Program run by USEPA.  The Contractor  can  expect
       to  analyze  at  least two Performance Evaluation (PE)  samples per  calendar
       quarter during  the  contract period  for organic low concentration
       analyses.

4.3    Annual  Verification of Method Detection Limits (MDLs)

       The  Contractor  shall perform  annual verification of MDLs in  accordance
       with the specifications in Exhibit  D.   All the MDLs shall  meet  the
       Contract Required Quantitation Limits  (CRQLs) specified in Exhibit C.

4.4    Quality Assurance/Quality Control Measurements

4.4.1      In this Exhibit, as well  as other places within this SOW, the term
           "analytical sample" is used in discussing the required frequency or
           placement of certain QA/QC measurements.   As  the term is  used,
           analytical sample  includes all field samples, including PE samples,
           received from an external  source.   It also includes all required
           QA/QC samples [requested Matrix Spike/Matrix Spike Duplicate(s)
            (MS/MSD), and Laboratory Control Sample  (LCS)]  except those  directly
           related to instrument calibration or calibration verification
            (calibration standards, Initial Calibration,  Continuing Calibration,
           and tunes).
 OLC03.2                                    E-8

-------
                                                              Exhibit E -- Sections 4 & 5
                                                                   Quality Assurance Plan

4.4.2     In order for the QA/QC information to reflect  the status of the
          samples analyzed, all samples and their QA/QC  analysis shall be
          analyzed under the same operating and procedural conditions.

4.4.3     If any QC measurement fails to meet contract criteria,-the analytical
          measurement must not be repeated prior to taking the appropriate
          corrective action as specified in Exhibit D.

4.4.4     The Contractor shall report all QC data in  the exact format specified
          in Exhibits B and H.

4.4.5     In addition, the Contractor shall establish a  QA program with the
          objective of providing sound analytical•chemical measurements.  This
          program shall incorporate the QC procedures, any necessary corrective
          action, and all documentation required during  data  collection, as
          well as the quality assessment measures performed by management to
          ensure acceptable data production.

5.0    QUALITY ASSURANCE PLAN (QAP)

5.1    Introduction

       The Contractor shall establish a Quality Assurance (QA) program with the
       objective of providing sound analytical chemical measurements.   This
       program shall incorporate the Quality Control  (QC) procedures,  any
       necessary corrective action,  all documentation required during data
       collection,  and the quality assessment measures performed by management
       to ensure acceptable data production.

5.1.1     As evidence of such a program, the Contractor  shall prepare'a written
          QAP which describes the procedures that are implemented to achieve
          the following:

          •    Maintain data integrity,  validity, and usability;

          •    Ensure that analytical measurement systems are maintained in an
               acceptable state of stability and reproducibility,-

          •    Detect problems through data assessment and establish corrective
               action procedures which keep the analytical process reliable,- and

          •    Document all aspects of the measurement process in order €o
               provide data which are technically sound and legally defensible.

5.1.2     The QAP must present, in  specific terms,  the policies, organization,
          objectives,  functional guidelines, and  specific QA/QC  activities
          designed to  achieve the data quality  requirements in this contract.
          Where  applicable,  Standard Operating  Procedures (SOPs) pertaining to
          each element  shall be included or referenced as part of the QAP.  The
          QAP shall be paginated consecutively  in ascending order.  The QAP
          shall  be available during on-site laboratory evaluations and shall  be
          submitted within  7 days of written request  by  the USEPA Regional
          Contract Laboratory Program Project Officer (CLP PO).  Additional
          information  relevant to the preparation of  a QAP can be found in
          USEPA  and American Society for Testing  and  Materials  (ASTM)
          publications.
                                            E-9                                     OLC03.2

-------
Exhibit E -- Section 5
Quality Assurance Plan (Con't)

5.2    Required Elements of a Quality Assurance Plan

       The required elements of a laboratory's QAP are outlined in this
       section.  This outline should be used as a framework for developing the
       QAP.

       A.   Organization  and Personnel

           1.   QA Policy and Objectives

           2.   QA Management

              a.  Organization

              b.  Assignment of QA/QC  Responsibilities

              c.  Reporting Relationships

              d.  QA Document Control  Procedures

              e.  QA Program Assessment Procedures

           3.    Personnel

              a.  Resumes

              b.  Education and Experience Pertinent to this Contract

              c.  Training Progress

       B.   Facilities and Equipment

           1.    Instrumentation and  Backup Alternatives

           2.    Maintenance Activities  and Schedules

       C.   Document Control

           1.    Laboratory Notebook  Policy

           2 .    Sample Tracking/Custody Procedures

           3.    Logbook Maintenance  and Archiving Procedures

           4.    Sample Delivery Group (SDG)  File  Organization,  Preparation,  and
                Review Procedures

           5.    Procedures for Preparation,  Approval,  Review,  Revision,  and
                Distribution of Standard Operating  Procedures (SOPs)

           6.    Process for Revision of Technical or  Documentation Procedures

       D.   Analytical Methodology

           1.    Calibration Procedures  and Frequency

           2.    Sample Preparation Procedures

           3 .    Sample Analysis Procedures

           4 .    Standards Preparation Procedures


 OLC03.2                                     E-10

-------
                                                                   Exhibit E -- Section 5
                                                           Quality Assurance Plan  (Con't)

          5.   Decision Processes,  Procedures,  and.Responsibility for Initiation
               of Corrective Action

       E.   Data  Generation

          1.   Data Collection Procedures

          2 .   Data Reduction Procedures

          3.   Data Validation Procedures    ;

          4.   Data Reporting and Authorization Procedures

       F.   Quality Assurance

          1.   Data Quality Assurance

          2.   Systems/Internal Audits

          3.   Performance/External Audits

          4 .   Corrective Action Procedures

          5.   QA Reporting Procedures

          6.   Responsibility Designation

       G.   Quality Control

          1.   Solvent, Reagent,  and Adsorbent Check Analysis

          2.   Reference Material Analysis

          3 .   Internal QC Checks

          4.   Corrective Action and Determination of QC Limit Procedures

          5.   Responsibility Designation

5.3    Updating and Submitting the Quality Assurance Plan

5.3.1     Initial  Submission.  During the  contract  solicitation process, the
          Contractor  is  required to  submit its QAP  to  the CLP Contracting
          Officer.  Within  sixty days after  contract award,  the Contractor
          shall  maintain on file a revised OAP.  fully  compliant with the
          recruirements of this contract.   The revised  QAP will become the
          official QAP under the contract  and may be used during  legal
          proceedings.   The Contractor shall maintain  the QAP on  file at the
          Contractor's facility  for  the  term of  the contract.  Both the initial
          submission  and the revised QAP shall be paginated  consecutively in
          ascending order.  The revised QAP shall include:

          •    Changes resulting from (1) the Contractor's internal review of
               their organization,  personnel, facility, equipment, policy and
               procedures and (2) the Contractor's implementation of the
               requirements of the contract;  and

          •    Changes resulting from USEPA's review of the laboratory
               evaluation sample data,  bidder supplied documentation,  and
               recommendations made during the pre-award on-site laboratory
               evaluation.                                               ^


                                           E-ll                                    OLC03.2

-------
Exhibit E -- Section 5
Quality Assurance Plan  (Con't)
5.3.1.1       The Contractor shall send'a copy of the latest version of the QAP
              within 7 days of a request from a USEPA Regional CLP PO.  The USEPA
              requestor will designate the recipients.

5.3.2     Subsequent  Updates and Submissions.   During the  term of  contract,  the
          Contractor  shall  amend the  QAP when  the  following circumstances
          occur:

          •    USEPA modifies the contract;f

          •    USEPA notifies the Contractor of deficiencies in the QAP
               document;

          •    USEPA notifies the Contractor of deficiencies resulting from
               USEPA's review of the Contractor's performance;

          •    The Contractor identifies  deficiencies  resulting from their
               internal review of their QAP document;

          •    The Contractor's  organization, personnel, facility, equipment,
               policy, or procedures  change; or

          •    The Contractor identifies  deficiencies  resulting from the
               internal review of their organization,  personnel, facility,
               equipment, policy, or  procedures changes.

5.3.2.1       The Contractor shall amend  the QAP within 30  days of when the
              circumstances  listed above  result in a discrepancy between what
              was previously described in the QAP and  what  is presently
              occurring at the Contractor's facility.  When the QAP is amended,
              all changes in the QAP  shall be clearly  marked (e.g., a bar in the
              margin  indicating  where the change is found  in the document, or
              highlighting the change by  underlining  the change, bold printing
              the change, or using a  different  print  font).  The amended section
              pages shall have the date on which the  changes were  implemented.
              The Contractor shall incorporate  all amendments to the latest
              version of the QAP document.  The Contractor shall archive all
              amendments to  the  QAP document for future reference  by USEPA.

5.3.2.2       The Contractor shall send  a copy  of the  latest version of the QAP
              document within 7  days  of  a written request  by the Regional CLP PO
              as directed.   The  USEPA requestor will  designate  the recipients.

5.4    Corrective Actions

       If the Contractor  fails  to adhere to the requirements listed  in this
       section,  the  Contractor  may expect,  but  USEPA is not limited  to the
       following actions:  reduction  in  the  numbers of samples  sent under  this
       contract,  suspension of  sample shipment  to  the Contractor,  data, package
       audit, electronic  data audit  (i.e.,  Gas  Chromatograph/Mass  Spectrometer
        (GC/MS)  tape  audit),  an  on-site  laboratory  evaluation,  remedial
       performance  evaluation sample, and/or contract sanctions.
 OLC03.2                                   E-12

-------
                                                                   Exhibit E -- Section 6
                                                            Standard Operating Procedures
6.0
6.1
       STANDARD OPERATING PROCEDURES (SOPs)
       Introduction
       In order to obtain reliable results,  adherence to prescribed analytical
       methodology is imperative.   In any operation that is performed on a
       repetitive basis,  reproducibility is  best accomplished through the use
       of Standard Operating Procedures (SOPs).   As defined by USEPA, a SOP is
       a written document which provides directions for the step-by-step
       execution of an operation,  analysis,  or action which is commonly
       accepted as the method for performing certain routine or repetitive
       tasks.

6.1.1     SOPs prepared by the Contractor shall be functional  (i.e., clear,
          comprehensive, up-to-date, and sufficiently detailed to permit
          duplication of results by qualified analysts).  The SOPs shall be
          paginated  consecutively  in ascending order.

6.1.2     All SOPs shall reflect activities as they  are currently performed  in
          the laboratory.  In addition, all SOPs shall be:              '

          •    Consistent with current USEPA regulations,  guidelines, and the
               Contract Laboratory Program (CLP)  contract's requirements.

          •    Consistent with instrument(s)  manufacturer's specific instruction
               manuals.

       : •  •    Available to USEPA during an on-site laboratory evaluation.   A
               complete set of SOPs shall be bound together and available for
      •i..       inspection at such evaluations.   During on-site laboratory
      f'       evaluations, laboratory personnel may be asked to demonstrate the
               application of the SOPs.
      '*."
          •    Available to the designated recipients within 7 days,  upon
       'i'       request by the USEPA Regional CLP Project Officer (CLP PO) .

      v  •    Capable of-providing for the development of documentation that is
               sufficiently complete to record the performance of all tasks
               required by the protocol.

          •    Capable of demonstrating the validity of data reported by the
               Contractor and explain the cause of missing or inconsistent
               results.

          •    Capable of describing the corrective measures and feedback
               mechanism utilized when analytical results do not meet protocol
               requirements.

          •    Reviewed regularly and updated as necessary when contract,
               facility,  or Contractor procedural modifications are made.

          •    Archived for future reference in usability or evidentiary
               situations.

          •    Available at specific work stations as appropriate.

          •    Subject to a document control procedure which precludes the use
               of outdated or inappropriate SOPs.
                                           E-13
                                                                                  OLC03.2

-------
Exhibit E -- Section 6
Standard Operating Procedures (Con't)

6.2    Format

       The format for SOPs may vary depending upon the type of activity for
       which they are prepared;  however,  at a minimum, the following sections
       shall be included:

       •    Title page;

       •    Scope and application;

       •    Definitions;

       •    Procedures;

       •    Quality Control (QC)  limits;

       •    Corrective action procedures,  including procedures for secondary
            review of information being generated;

       •    Documentation description  and example forms;

       •    Miscellaneous notes  and precautions;  and

       •    References.

 6.3    Requirements

       The  Contractor shall maintain the  following SOPs:

 6.3.1      Evidentiary SOPs for  required  Chain-of-Custody and document control
           are discussed in Exhibit F.

 6.3.2      Sample Receipt and Storage

           •   Sample receipt and  identification logbooks;

           •   Refrigerator temperature logbooks;  and

           •   Security  precautions.

 6.3.3      Sample Preparation

           •   Reagent purity check procedures and documentation;

           •   Extraction procedures;

           •   Extraction bench sheets,- and

           •   Extraction logbook  maintenance.

 6.3.4      Glassware Cleaning

 6.3.5      Calibration (Balances, etc.)

           •   Procedures;

           •   Frequency requirements;

           •   Preventive maintenance  schedule and procedures;  and

           •   Acceptance criteria and corrective  actions.


 OLC03.2                                    E-14

-------
                                                                   Exhibit E  -- Section 6
                                                    Standard Operating Procedures  (Con't)

6.3.6     Analytical Procedures  (for each analytical  system)

          •    Instrument performance specifications;

          •    Instrument operating procedures;

          •    Data acquisition system operation;

          •    Procedures when automatic quantitation algorithms are overridden;

          •    QC required parameters;

          •    Analytical run/injection logbooks; and

          •    Instrument error and editing.flag descriptions and resulting
               corrective actions.

6.3.7     Maintenance  Activities  (for  each  analytical system)

          •    Preventative maintenance schedule and procedures;

          •    Corrective maintenance determinants and procedures; and

          •    Maintenance authorization.

6.3.8     Analytical Standards

          •    Standard coding/identification and inventory system;

          •    Standards preparation logbook(s);

          •    Standard preparation procedures;

          •    Procedures for equivalency/traceability analyses and
               documentation;

          •    Purity  logbook (primary standards and  solvents);

          •    Storage, replacement, and labeling requirements; and

          •    QC and  corrective action measures.

6.3.9     Data Reduction Procedures

          •    Data processing systems operation;

          •    Outlier identification methods;

          •    Identification of data requiring corrective action; and

          •    Procedures for format and/or forms for each operation.

6.3.10    Documentation Policy/Procedures

          •    Contractor/analyst's notebook policy,  including review policy;

          •    Complete SDG File contents;

          •    Complete SDG File organization and assembly procedures, including
               review  policy; and

          •    Document inventory procedures,  including review policy.

                                           E-15                                    OLC03.2

-------
Exhibit E -- Section 6
Standard Operating Procedures  (Con't)


6.3.11    Data Validation/Self-Inspection Procedures

          •    Data flow and chain-of-command for data review;

          •    Procedures for measuring precision and accuracy;

          •    Evaluation parameters for  identifying systematic -errors;

          •    Procedures to assure that  hardcopy and electronic deliverables
               are complete and compliant with the requirements in the Statement
               of Work  (SOW) Exhibits B and H;

           •    Procedures to assure that  hardcopy deliverables are in agreement
               with their comparable electronic deliverables;

           •    Demonstration of internal  QA inspection procedure  (demonstrated
               by supervisory  sign-off on personal notebooks, internal
                laboratory evaluation samples, etc.);

           •    Frequency and type of internal audits  (e.g.,  random, quarterly,
                spot checks, perceived trouble areas) ,-

           •    Demonstration of problem identification,  corrective actions,  and
                resumption of analytical processing.  Sequence resulting  from
                internal audit  (i.e., QA feedback); and

           •     Documentation of audit reports  (internal  and external), response,
                corrective action, etc.

 6.3.12     Data Management and Handling

           •     Procedures for  controlling and estimating data entry errors,-

           •    Procedures for  reviewing changes to data  and deliverables and
                ensuring traceability of updates;

           •    Lifecycle management  procedures  for testing,  modifying, and
                implementing changes  to existing computing systems  including
                hardware,  software, and documentation or  installing new systems;

           •    Database security, backup, and archival procedures  including
                recovery from system  failures;

           •    System maintenance procedures and  response time;

           •    Individual(s)  responsible  for  system  operation,  maintenance,  data
                integrity,  and security,-

           •    Specifications for staff  training  procedures; and

           •    Virus protection procedures for  software  and electronic
                deliverables.

 6.4    Updating and Submitting SOPs

 6.4.1     Initial Submission.  During the contract  solicitation process,  the
           Contractor  is required to submit their SOPs to the CLP Contracting
           Officer.  Within 60 days after contract award,  the Contractor shall
           maintain on file a  complete revised set of SOPs.  fully compliant  with
           the requirements of this contract.  The revised SOPs will become  the
           official SOPs under the contract and may be used  during legal

 OLC03.2                                    E-16

-------
                                                                   Exhibit E -- Section 6
                                                    Standard Operating Procedures (Con't)

          proceedings.  The Contractor shall maintain the complete set of SOPs
          on file at the Contractor's facility for the term of the contract.
          Both the initial submission of SOPs and the revised SOPs shall be
          paginated consecutively in ascending order.  The revised SOPs shall
          include:

          •    Changes resulting from 1)  the Contractor's internal review of
               their procedures,  and 2)  the Contractor's implementation of the
               requirements of the contract,  and

          •    Changes resulting from USEPA's review of the laboratory
               evaluation sample data,  bidder supplied documentation,  and
               recommendations made during the preaward on-site laboratory
               evaluation.

6.4.1.1       The Contractor shall send a complete  set of the  latest version of
              SOPs or individually requested SOPs within 7 days of a request
              from an USEPA Regional CLP PO.   The USEPA requestor will designate
              the recipients.

6.4.2     Subsequent Updates and Submissions.  During the term of the contract,
          the Contractor shall amend the SOPs when the following circumstances
          occur:

          •    USEPA modifies the contract;

          •    USEPA notifies the Contractor of deficiencies in their  SOP's
               documentation;

          •    USEPA notifies the Contractor of deficiencies resulting from
               USEPA's review of the Contractor's performance;

          •    The Contractor's procedures change;

          •    The Contractor identifies deficiencies resulting from the
               internal review of their SOPs documentation; or

          •    The Contractor identifies deficiencies resulting from the
               internal review of their procedures.

6.4.2.1       Existing SOPs shall be amended or new SOPs shall be written within
              30  days of when the circumstances listed above result  in a
              discrepancy between what was previously described in the SOPs and
              what is presently occurring at the Contractor's  facility.   All
              changes in the SOPs shall be clearly  marked (e.g.,  a bar in the
              margin indicating where the change is in the document, or
              highlighting the change by underlining the change,  bold  printing
              the change,  or using a different print font).   The amended/new
              SOPs shall have the date on which the changes were implemented.

6.4.2.2       When existing SOPs are amended or new SOPs are written,  the
              Contractor shall document the reasons for the changes  and maintain
              the amended SOPs or new SOPs on file.   Documentation of  the
              reasons for the changes shall be maintained on file with the
              amended SOPs or new SOPs.

6.4.2.3       Documentation of the reason(s)  for changes to the SOPs shall also
              be  submitted along with the SOPs.   An alternate  delivery schedule
              for submitting the letter and amended/new SOPs may be proposed by
              the Contractor, but it is the sole decision of the USEPA
              Contracting Officer to approve or disapprove the alternate
              delivery schedule.   If an alternate delivery schedule  is proposed,

                                           E-17                                   OLC03.2

-------
Exhibit E -- Section 6
Standard Operating Procedures  (Con't)

              the Contractor shall describe in a letter to the USEPA Regional
              CLP PO and the Contracting Officer why it is unable to meet the
              delivery schedule listed in this section.  The USEPA Regional CLP
              PO may grant an extension of up to 30 days for amending/writing
              new SOPs.  An extension for amending/writing new SOPs beyond 30
              days must be approved by the USEPA Contracting Officer.
              Similarly, an extension of up to 14 days for submission of the
              letter documenting the reasons for the changes and for submitting
              amended/new SOPs may be approved by the USEPA Regional CLP PO.  An
              extension beyond the 14 days must be approved by the USEPA
            -  Contracting Officer.  The Contractor shall proceed and not assume
              that an extension will be granted until so notified by the USEPA
              Regional CLP PO and/or Contracting Officer.

6.5    Corrective Action

       If the Contractor fails to adhere to the requirements listed in this
       section, the Contractor may expect, but USEPA is not limited to the
       following actions: reduction  in the number of samples sent under this
       contract, suspension of sample shipment to the Contractor, data package
       audit, electronic data audit, an on-site laboratory evaluation, remedial
       performance evaluation sample, and/or contract sanctions.
 OLC03.2                                    E-18

-------
                                                                   Exhibit E  -- Section 7
                                                            Contract Compliance Screening

7.0    CONTRACT COMPLIANCE SCREENING  (CCS)

7.1    Overview

7.1.1      CCS  is  one aspect  of the Government's  contractual right  of inspection
           of analytical data.   CCS examines  the  Contractor's  adherence to the
           contract  requirements based on the Sample Data  Package delivered to
           USEPA.

7.1.2      CCS  is  performed by the Sample Management Office  (SMO) under the
           direction of USEPA.   To assure a uniform review, a  set of
           standardized procedures has been developed  to evaluate the Sample
           Data Package submitted by a Contractor against  the  technical and
           completeness requirements of the contract.  USEPA reserves the right
           to add  and/or delete individual checks.

7.2    CCS Results

       CCS results are  mailed  to  the  Contractor and all other data recipients.
       The Contractor has a period  of  time to correct deficiencies.  The
       Contractor shall send all  corrections to the Regional client and SMO
       within  6 business days.  CCS results are used in conjunction with other
       information to measure  overall  Contractor performance and to take
       appropriate actions to  correct  deficiencies in performance.

.7.3    CCS Trend Report

       USEPA may generate a CCS trend report which summarizes CCS results over
       a given period of time.  USEPA may send the CCS trend report or discuss
       the CCS trend report during  an on-site laboratory evaluation.   In a
       detailed letter  to the  USEPA Regional Contract Laboratory Program
       Project Officer  (CLP PO) and Contracting Officer,  the Contractor shall
       address the deficiencies and the  subsequent corrective action
       implemented by the Contractor  to  correct the deficiencies within 14 days
       of receipt of the report of  the on-site laboratory evaluation.  An
       alternate delivery schedule  may be proposed by the Contractor, but it is
       the sole decision of USEPA to  approve or disapprove the alternate
       delivery schedule.  If  an  alternate delivery schedule is proposed, the
       Contractor shall describe  in a letter to the USEPA Regional CLP PO and
       Contracting Officer why it is  unable to meet the delivery schedule
       listed  in this section.  The USEPA Regional CLP PO may grant an
       extension of up  to 14 days for the Contractor's response to the CCS
       trend report.  An extension  beyond 14 days must be approved by the USEPA
       Contracting Officer.  The  Contractor shall proceed and not assume that
       an extension will be granted until so notified by the appropriate USEPA
       official.
7.4    Corrective Actions

7.4.1      If  new Standard Operating Procedures  (SOPs)  are  required to be
           written,  or if existing SOPs are required  to be  rewritten or amended
           because of deficiencies and subsequent  corrective  action implemented
           by  the Contractor,  the Contractor shall write/amend the SOPs per the
           requirements listed in Section 6.

7.4.2      If  the Contractor fails to adhere to  the requirements  listed in this
           section,  the Contractor may expect, but USEPA  is not limited to the
           following actions:  reduction in the number of  samples  sent under the
           contract,  suspension of sample shipment to the Contractor,  data
           package audit,  electronic data audit, an on-site laboratory
           evaluation,  a remedial performance evaluation  sample,  and/or contract
           sanctions.

                                           E-19
                                                                                  OLC03.2

-------
Exhibit E -- Section 8
Analytical Standards Requirements

8.0    ANALYTICAL STANDARDS REQUIREMENTS

8.1    Overview

       USEPA may not supply analytical reference standards either for direct
       analytical measurements or for the purpose of traceability.   All .
       contract laboratories shall be required to prepare from materials or
       purchase from private chemical supply houses those standards necessary
       to successfully and accurately perform the analyses required in this
       protocol.

8.2    Preparation of Chemical Standards from the Neat High Purity Bulk
       Material

8.2.1      If  the  laboratory cannot  obtain  analytical  reference standards,  the
           laboratory may prepare its own chemical  standards.  Laboratories
           shall obtain the highest  purity  possible when  purchasing  chemical
           standards;  standards purchased at  less than 97% purity shall  be
           documented as to why a higher purity could  not be obtained.

8.2.2      If  required by the manufacturer, the chemical  standards shall be kept
           refrigerated when not being used in the  preparation of standard
           solutions.   Proper storage of chemicals  is  essential in order to
           safeguard them from decomposition.

8.2.3      The purity of a compound  can sometimes be misrepresented  by a
           chemical supply house. Since knowledge  of  purity is needed to
           calculate the concentration of solute in a  solution standard, it is
           the Contractor's responsibility  to have  analytical  documentation
           ascertaining that the purity of  each compound  is  correctly stated.
           Purity confirmation, when performed,  should use appropriate
           techniques.  Use of two or more  independent methods is recommended.
           The correction factor for impurity when  weighing  neat  materials  in
           the preparation of solution standards is:

           EQ. 1
                 weight ot impure expound -  reiobt ot pure
                                                  (percent pur icy/100)
           Where: "weight of pure compound" is that required to prepare a
           specific volume of a solution standard of a specified concentration.

 8.2.4     When compound purity is assayed to be 97 percent or greater,  the
           weight may be used without correction to calculate the concentration
           of the stock standard.  If the compound purity is assayed to be less
           than 97 percent, the weight shall be corrected when calculating the
           concentration of the stock solution.

 8.2.5     Mis-identification of compounds occasionally occurs and it is
           possible that a mislabeled comp'ound may be received from a chemical
           supply house.  It is the Contractor's responsibility to have
           analytical documentation ascertaining that all compounds used in the
           preparation of solution standards are correctly identified.
           Identification confirmation, when performed, shall use Gas
           Chromatography/Mass Spectrometry (GC/MS) analysis on at least two
           different analytical columns, or other appropriate techniques.

 8.2.6     Calculate the weight of material to be weighed out for a specified
           volume, taking into account the purity of the compound and the

 OLC03.2                                   E-20

-------
                                                                    Exhibit E -- Section 8
                                                 Analytical Standards Requirements (Con't)

           desired concentration.  A second person shall verify the accuracy of
           the calculations.  Check balances for accuracy with a set of standard
           weights every 12 hours.  All weighing shall be performed on an
           analytical balance to the nearest 0.1 milligram (mg) and verified by
           a second person.  The solvent used to dissolve the solute shall be
           compatible with the protocol in which the standard is to be used; the
           solute shall be soluble, stable, and non-reactive with the solvent.
           In the case of a multicomponent solution, the components must not
           react with each other.

 8.2.7     Transfer the solute to a volumetric, flask and dilute to the specified
           solution volume with solvent after ensuring dissolution of the solute
           in the solvent.  Sonication or warming may be performed to promote
           dissolution of the solute.  This solution shall be called the primary
           standard and all subsequent dilutions shall be traceable back to the
           primary standard.

 8.2.8     Log notebooks are to be kept for all weighing and dilutions.  All
           subsequent dilutions from the primary standard and the calculations
           for determining their concentrations are to be recorded and verified
           by a second person.  All solution standards are to be refrigerated,
           if required, when not in use.  All solution standards are to be
           clearly labeled as to the identity of the analyte or analytes,
           concentration, date prepared, solvent, and initials of the preparer.

'•'8.3    Purchase of Chemical Standards Already in Solution

        Solutions of analytical  reference standards can be purchased by
        Contractors provided they meet the following criteria.

 8.3.1 ...   Contractors shall maintain documentation of the purity confirmation
       .>:',   of the material to verify the integrity of the standard solutions
       =•*!   they purchase.

 8.3.2     The Contractor shall purchase standards for which the quality is
           demonstrated statistically and analytically by a method of the
           supplier's choice.  One way this can be demonstrated is to prepare
           and analyze three solutions: a high standard, a low standard, and a
           standard at the target concentration  (Sections 8.3.2.1 and 8.3.2.2).
           The supplier must then demonstrate that the analytical results for
           the high standard and low standard are consistent with the difference
           in theoretical concentrations.  This is done by the student's t-test
           in Section 8.3.2.4.   If this is achieved, the supplier must then
           demonstrate that the concentration of the target standard lies midway
           between the concentrations of the low and high standards.  This is
           done by the student's t-test in Section 8.3.2.5.  Thus the standard
           is certified to be within 10 percent of the target concentration
           using the equations in Section 8.3.2.6.  If the procedure described
           above is used, the supplier must document that the following have
           been achieved.

 8.3.2.1       Two solutions of  identical  concentration shall be prepared
               independently from neat materials.   An aliquot of the first
               solution shall be diluted to the  intended concentration (the
               "target  standard").   One aliquot  is taken from the  second solution
               and diluted to a  concentration 10  percent greater than the target
               standard.   This is called the "high standard".   One further
               aliquot  is taken  from the second  solution and diluted to a
               concentration 10  percent less than the target standard.   This  is
               called the "low standard".
                                            E-21                                   OLC03.2

-------
Exhibit E -- Section 8
Analytical Standards Requirements (Con't)

8.3.2.2       Six replicate analyses of  each standard (a total  of  18  analyses)
              shall be performed in  the  following sequence:  low standard,  target
              standard,  high standard; low standard,  target  standard,  high
              standard;  etc.

8.3.2.3       The mean and variance  of the six results for each solution shall
              be calculated:

              EQ. 2

                                   Y. + Y. + Y, + Y. + YK +  Y.
                           BEAM -  *     *    *      •   *    *
              EQ. 3

                             Y* + Y*  + Y* + Y* + Y* + Y* - 6(HE1N>*
                 VARIANCE • =	=	 *    4     *     *
              The values YK Y2, Y3,  ...,  represent the results  of  the six
              analyses of each standard.   The  means of the  low, target, and high
              standards are designated Mlr M2,  and M3,  respectively.  The
              variances of the low,  target,  and high.standards  are designated
              Vlf  V2, and V3, respectively.  Additionally, a pooled variance, Vp,
              is calculated.

              EQ. 4
              If the square root of Vp is less than one percent of M2,  then
              M22/10,000 is to be used as the value of Vp in all  subsequent
              calculations.

8.3.2.4       The test statistic shall be  calculated:

              EQ. 5
                                                _!».-  A
                                                 1.1    0.9
                                                    „,...
                            TEST STATISTIC •
              If  the  test  statistic exceeds 2.13, then the supplier has failed
              to  demonstrate  a  20 percent difference between the high and low
              standards.   In  such a case, the  standards are not acceptable.
 OLC03.2                                   E-22

-------
                                                                   Exhibit E -- Section 8
                                                Analytical Standards Requirements (Con't)
                                                           3_
                                                           2.2
8.3.2.5       The  test statistic shall be calculated:

              EQ.  6
                       TEST STATISTIC -
              If  the test  statistic exceeds 2.13, the supplier has failed to
              demonstrate  that  the target standard concentration is midway
              between  the  high  and low standards.  In such a case, the standards
              are not  acceptable.

 8.3.2.6      The 95 percent confidence intervals for the mean result of each
              standard shall be calculated:

              EQ. 7


                     Interval tor Low Standard - H, ±2.13
              EQ. B


                   Interval tor Target Standard - Hj ± 2.13



              EQ. 9


                     Interval  tor High Standard - H, ± 2.13


 8.3.2.6.1        These intervals shall not overlap.  If overlap is observed,
                  then the supplier has failed  to demonstrate the ability to
                  discriminate  the 10  percent difference in concentrations.  In
                  such a case,  the standards are not acceptable.

 8.3.2.6.2        In  any event,  the Contractor  is responsible for the quality of
                  the standards employed  for analyses under this contract.

 8.4    Requesting Standards from the USEPA Standards Repository

        Solutions  of  analytical  reference materials can be ordered from the
        USEPA Chemical  Standards Repository, depending on availability.  • The
        Contractor may  place an  order for standards only after demonstrating
        that  these standards are not available  from commercial vendors,  either
        in solution or  as  a neat material.

 8.5    Documentation of the Verification and Preparation of Chemical Standards

        It is the  responsibility of the Contractor to maintain the necessary
        documentation to show that the chemical standards it has used in the
        performance of  Contract  Laboratory Program (CLP) analysis conform to the
        requirements  previously  listed.


                                           E-23                                   OLC03.2

-------
Exhibit E -- Section 8
Analytical Standards Requirements  (Con't)

8.5.1     Weighing  logbooks, calculations, raw data, etc., whether produced by
          the Contractor or purchased from chemical supply houses, shall be
          maintained by the Contractor and may be subject to review during on-
          site  inspection visits.   In those cases where the documentation is
          supportive of the analytical results of data packages sent to USEPA,
          such  documentation is  to be kept on file by the Contractor for a
          period  of one year.

8.5.2     Upon  request by the  USEPA Regional CLP Project Officer  (CLP  PO), the
          Contractor shall submit  their  most recent previous year's
          documentation  (12 months) for  the verification and preparation of
          chemical  standards within 14 days of the receipt of  request  to the
          designated recipients.

8.5.3     USEPA may generate a report discussing deficiencies  in  the
          Contractor's documentation for the verification and  preparation of
          chemical  standards or  may discuss the deficiencies during an on-site
          laboratory evaluation.   In a detailed letter to the  USEPA Regional
          CLP PO  and Quality Assurance Technical Support  (QATS),  the Contractor
          shall address the deficiencies and the subsequent corrective action
          implemented by the Contractor  to correct the deficiencies within 14
          days  of receipt of the report  or the on-site laboratory evaluation.
          An alternate delivery  schedule may be proposed by the Contractor, but
          it is the sole decision of USEPA to approve or disapprove the
          alternate delivery schedule.   If an alternate delivery  schedule is
          proposed, the Contractor shall describe  in a letter  to  the USEPA
          Regional  CLP PO and  the Contracting Officer why it is unable to meet
          the delivery schedule  listed  in this section.  The USEPA Regional CLP
          PO may  grant an extension of up to  14 days for  the Contractor's
          response letter to the standards documentation  report.  An extension
          beyond  14 days must  be approved by  the USEPA Contracting Officer.
          The  Contractor  shall proceed  and not assume that an  extension will  be
          granted until  so  notified by  the appropriate USEPA official.

8.5.4      If new  Standard Operating Procedures  (SOPs) are required to  be
          written,  or if  existing SOPs  are required to be rewritten or amended
          because of deficiencies and subsequent, corrective action implemented
          by the  Contractor,  the Contractor  shall  write/amend  the SOPs per the
           requirements  listed  in Section 6.

8.6     Corrective Actions

        If the Contractor fails to  adhere to the requirements listed in this
        section, a Contractor may expect,  but USEPA is not limited to the
        following  actions: reduction in the number of samples sent under the
        contract,  suspension of sample shipment to the Contractor, data package
        audit, electronic data  audit, an  on-site laboratory evaluation, a
        remedial laboratory evaluation sample, and/or contract  sanctions.
 OLC03.2-                                   E-24

-------
                                                                    Exhibit  E  --  Section  9
                                                                       Data  Package Audits
9.0
       DATA PACKAGE AUDITS
9.1    Overview

       Data package audits are performed by USEPA for program overview and
       specific Regional concerns.  Standardized procedures have been
       established to assure uniformity of the auditing process.  Data packages
       are periodically selected from recently received Cases.  They are
       evaluated for the technical quality of hardcopy raw data, Quality
       Assurance (QA),  and the adherence to contractual requirements.  This
       function provides external monitoring of program Quality Control (QC)
       requirements.   Data package audits are used to assess the technical
       quality of the data and evaluate overall laboratory performance.  Audits
       provide USEPA with an in-depth inspection and evaluation of the Case
       data package with regard to achieving QA/QC acceptability.  A-thorough
       review of the raw data is completed including: all instrument readouts
       used for the sample results, instrument printouts, and other
       documentation for deviations from the contractual requirements, a check
       for transcription and calculation errors, a review of the qualifications
       of the laboratory personnel involved with the Case, and a review of the
       latest version of all Standard Operating Procedures (SOPs) on file.

9.2    Responding to the Data Package Audit Report

9.2.1     After completion  of  the data  package  audit, USEPA  may send a copy of
:—      ;,.  the  data package  audit  report to the  Contractor or may discuss  the
          data package  audit report  on  an on-site  laboratory evaluation.   In a
          detailed letter to the  USEPA  Regional Contract Laboratory Program
          Project Officer (CLP PO) and  the USEPA designated  recipient,  the
 •'•• :    •.'•;  Contractor shall  discuss the  corrective  actions implemented to
      -r  resolve the deficiencies listed in the data package audit report
• -     ••";•  within 14  days  of receipt  of  the report.
 - - -    f\
9.2.2     An alternate  delivery schedule may be proposed by  the Contractor,  but
 -'•:'':    ."•,  it is the  sole  decision of USEPA to approve or disapprove the
 "'-"•     ;•  alternate  delivery schedule.   If an alternate  delivery schedule is
~^::.     -j~  proposed,  the Contractor shall describe  in a letter to the USEPA
          Regional CLP  PO and  the Contracting Officer why it is unable to meet
          the  delivery  schedule listed  in this  section.   The USEPA Regional CLP
          PO may grant  an extension  of  up to 14 days for the Contractor's
          response letter to the  data package report.  An extension beyond 14
          days must  be  approved by the  USEPA Contracting Officer.   The
          Contractor shall  proceed and  not assume  that an extension will  be
          granted until so  notified  by  the appropriate USEPA official.

9.2.3     If new SOPs are required to be written,  or if  existing SOPs are
          required to be  rewritten or amended because of deficiencies and
          subsequent corrective action  implemented by the Contractor,  the
          Contractor shall  write/amend  the SOPs per the  requirements listed in
          Section 6.

9.3    Corrective Actions

       If the Contractor fails to adhere to the requirements listed in this
       section,  the Contractor may expect, but USEPA is not limited to the
       following actions: reduction in  the number of samples sent under the
       contract, suspension of sample shipment to the Contractor, an.on-site
       laboratory evaluation, data package audit, electronic data audit,
       remedial -performance evaluation  sample, and/or contract sanctions.
                                           E-25
                                                                                   OLC03.2

-------
Exhibit E -- Sections 10 & 11
Regional Data Review

10.0   REGIONAL DATA REVIEW

10.1   Overview.

       Contractor data are generated to meet the specific needs of USEPA
       Regions.  In order to verify the usability of data for the intended
       purpose, each Region reviews data from the perspective of the end user,
       based on functional guidelines for data review which have been developed
       jointly by the Regions and the National Program Office.  Each Region
       uses these guidelines as the basis for data evaluation.  Individual
       Regions may augment the basic guideline review process with additional
    1   review based on Region-specific or site-specific concerns.  Regional
       reviews, like the sites under investigation, vary based on the nature of
       the problem under investigation and the Regional response appropriate to
       the specific circumstances.

10.1.1    Regional  data  reviews,  relating  usability of  the  data  to  a  specific
          site,  are part of  the  collective assessment process.   They  complement
          the  review performed at  the  Sample Management Office  (SMO),  which is
          designed  to identify contractual discrepancies, and the review
          performed by the National  Program Office,  which is  designed to
          evaluate  Contractor and method performance.   These  individual
          evaluations are integrated into  a collective  review that  is necessary
          for  Program and Contractor administration and management, and  may be
          used to take appropriate action  to correct deficiencies in  the
          Contractor's performance.

11.0   PROFICIENCY TESTING

       As a means of measuring and evaluating both the  Contractor's and  the
       method's analytical performance,  the  Contractor  must participate  in
       USEPA's  Proficiency Testing Program.  USEPA's Proficiency Testing
       Program involves  the  analysis of  Case specific Performance Evaluation
        (PE) samples  and  the  participation  in interlaboratory Quarterly Blind
        (QB) Audits.  The Contractor's  analytical PE  samples and QB results will
       be used by USEPA  to assess and  verify the Contractor's continuing
       ability to produce acceptable analytical data in accordance with  the
       contractual  requirements.

11.1   Performance  Evaluation (PE) Samples

11.1.1     The  PE sample(s)  may be scheduled with the Contractor as  frequently
           as on a Sample Delivery Group (SDG)-by-SDG basis.  The PE samples may
           be sent either by the  Regional client or the National Program Office.
           PE samples will assist USEPA in monitoring Contractor performance.

11.1.2     PE samples will be provided as either single-blinds (recognizable as
           a PE sample but of unknown composition), or  as double-blinds  (not
           recognizable as a PE sample and of unknown composition).   The
           Contractor will not be informed of either the analytes/parameters or
           the concentrations in the PE samples.

11.1.3     The Contractor may receive the PE samples as either full  volume
           samples or ampulated/bottled concentrates from USEPA or a designated
           USEPA Contractor.   The PE samples shall come with instructions
           concerning the unique preparation procedures, if  any,  required to
           reconstitute the PE samples (i.e., the required dilution of the PE
           sample concentrate).  PE samples are to be prepared and analyzed with
           the rest of the routine samples in the SDG.   The  Contractor shall
           prepare and analyze the PE sample using the  procedure described in
           the sample preparation and method analysis sections of Exhibit D.
           All contract required Quality Control (QC) shall  also be met.   The PE

OLC03.2                                    E-26

-------
                                                                   Exhibit E -- Section 11
                                                               Proficiency Testing (Con't)

           sample  results  are  to  be  submitted in the SDG deliverable package per
           normal  reporting  procedures  detailed in Exhibit B.

11.1.4     In addition to  PE sample  preparation and analysis,  the Contractor
           shall be  responsible for  correctly identifying and quantitating the'
           analytes/parameters included in each PE sample.  When PE sample
           results are received by .USEPA,  the PE sample results will be
           evaluated for correct  'analytical identification and quantitation.
           The  PE  sample evaluation  will be provided to the Contractor via coded
           evaluation sheets,  by  analyte/parameter.   USEPA will notify the
           Contractor of unacceptable performance.  USEPA reserves the right to
           adjust  the PE sample acceptance windows in order to compensate for
           any  unanticipated difficulties with a particular PE sample.

11.1.5     The  Contractor  shall demonstrate acceptable analytical performance
           for  both  identification and  quantitation of PE sample analytes/
           parameters.  For  unacceptable PE sample performance, USEPA may take,
           but  is  not limited  to,-the following actions: reduce value or
           rejection of data for  the samples, SDG, or Case impacted, contract
           sanctions, reduction in the  number of samples shipped to the
           laboratory, suspension of sample shipment, an on-site laboratory
           inspection, a full  data package audit, electronic data audit,  and/or
           require the laboratory to analyze a Remedial QB.

           NOTE: A Contractor's prompt  response demonstrating that corrective
           actions have been taken to ensure the Contractor's capability to meet
           contract  requirements  may facilitate continuation of full sample
           delivery.

11.2   .'Quarterly  Blind (QB) Audits

ill.2.1-'£   QB Audits may be  scheduled concurrently with all contract
:-".    -:.=;    .:~:   office  (SMO) .  QB samples will assist USEPA in monitoring Contractor
:n:-.    \:   performance.

11.2.2     QB samples will be  provided  as single-blinds (recognizable as a PE
           sample  but of unknown  composition).  The Contractor will not be
           informed  of either  the analytes or the concentrations in the PE
           samples.

11.2.3     The  Contractor  may  receive the QB samples as either full volume
           samples or ampulated/bottled concentrates from USEPA or a designated
           USEPA Contractor.  The QB samples shall come with instructions
           concerning the  unique  preparation procedures, if any, required to
           reconstitute the  QB samples  (i.e., the required dilution of the QB
           sample  concentrate).   The Contractor shall prepare and analyze the QB
           samples using the procedure  described in the sample preparati.on and
           method  analysis sections  of  Exhibit D.  All contract required QC
           shall also be met.  The QB sample results are to be submitted in the
           SDG  deliverable package per  normal reporting procedures detailed in
           Exhibit B.

11.2.4     In addition to  QB sample  preparation and analysis,  the Contractor
           shall be  responsible for  correctly identifying and quantitating the
           analytes/parameters included in each QB sample.  When QB sample
           results are received by USEPA, the QB sample results will be scored
           for  correct analytical identification and quantitation.  The QB
           sample  scoring  will be provided to the Contractor via coded
           evaluation sheets,  by  analyte/parameter.   USEPA wil-1 notify the

                                           E-27                                    OLC03.2

-------
Exhibit E -- Section 11
Proficiency Testing (Con't)

          Contractor of .unacceptable performance.  USEPA reserves the right to
          adjust the PE sample acceptance windows in order to compensate for
          any unanticipated difficulties with a particular PE sample.  The
          Contractor's QB  sample performance will be assessed into one of the
          following three  categories:

11.2.4.1      Acceptable,  No Response Required:  Score greater than or equal to
              90 percent.   The data meets most or all of the scoring criteria.
              No response is required.

11.2.4.2      Acceptable,  Response Explaining Deficiencies Required:  Score
              greater than or equal to 75 percent,  but less than 90 percent.
              Deficiencies exist in the Contractor's performance.  Corrective
              action response required.

11.2.4.3      Unacceptable Performance, Response Explaining Deficiencies
              Required: Score less than 75 percent.   Deficiencies exist in the
              Contractor's performance to the extent that the National Program
              Office has determined that the Contractor has not demonstrated the
              capability to meet the contract requirements.  Corrective action
              response required.

11.2.5    In the case  of  Section 11.2.4.2  or 11.2.4.3,  the Contractor shall
          describe the deficiency(ies)  and the  action(s)  taken  in a  corrective
          action letter to the USEPA Regional Contract  Laboratory Program
          Project Officer (CLP PO)  and CLP Quality  Assurance (QA)  Coordinator
          within 14  days  of receipt of notification from USEPA.

11.2.5.1      An alternate delivery schedule for the corrective action letter
              may .be proposed by the Contractor, but it is the sole decision of
              USEPA to approve or disapprove the alternate delivery schedule.
              If an alternate delivery schedule is proposed, the Contractor
              shall describe in a letter to the USEPA Regional CLP PO and
              Contracting Officer why the laboratory is unable to meet the
              original delivery schedule listed in Section 11.2.5.  The USEPA
              Regional CLP PO may grant an extension of up to 14 days for the
              Contractor's corrective action .letter.  An extension beyond 14
              days must be approved by the USEPA Contracting Officer.  The
              Contractor shall proceed and not assume that an extension will be
              granted until so notified by the appropriate USEPA official.

11.2.6     In the case of Section 11.2.4.2 or 11.2.4.3,  if new Standard
           Operating Procedures (SOPs)  are required to be written,  or if
           existing SOPs are required to be rewritten or amended because of
           deficiencies and subsequent  corrective action implemented by the
           Contractor,  the Contractor shall write/amend the SOPs per the
           requirements listed in Section 6.

11.2.7     The Contractor shall be notified by the USEPA Contracting Officer
           concerning agreement or disagreement with the proposed remedy for
           unacceptable performance.  For unacceptable QB sample performance
           (Section 11.2.4.3), USEPA may take, but is not limited to, the
           following actions: reduction in the number of samples shipped to the
           laboratory,  suspension of sample shipment, an on-site laboratory
           inspection,  electronic data audit, a full data package audit, and/or
           require the laboratory to analyze a Remedial QB sample, and/or
           contract sanctions.

           NOTE: A Contractor's prompt response demonstrating that corrective
           actions have been taken to ensure the Contractor's capability to meet
           contract requirements may facilitate continuation of full sample
           delivery.

 OLC03.2                          .         E-28

-------
                                                                  Exhibit E -- Section 11
                                                              Proficiency Testing  (Con't)
11.2.8    A Remedial QB Audit is a unique analytical Case containing only QB
          samples.  A Remedial QB Audit may be scheduled by the National
          Program Office with the Contractor(s) for any of the following
          reasons: unacceptable PE sample performance, unacceptable QB sample
          performance, arid/or major change in the laboratory  (e.g., relocation,
          new owner, or high turnover of key personnel).  Sections 11.2.2
          through 11.2.7 apply to the Remedial QB Audit process.

11.3   Corrective Actions

       If the Contractor fails to adhere to the requirements lasted in this
       section,  the Contractor may expect,  but USEPA is not limited to,  the
       following actions:  reduction in the number of samples sent under the
       contract, suspension of sample shipment to the Contractor,»a full data
       package audit,  electronic data audit,  an on-site laboratory inspection,
       a Remedial QB sample, and/or contract sanctions.
                                           E-29                                   OLC03.2

-------
Exhibit E -- Section 12
On-Site Laboratory Evaluations

12.0   ON-SITE LABORATORY EVALUATIONS

12.1   Overview

       As dictated by a contract laboratory's performance,  the USEPA Regional
       Contract Laboratory Program Project Officer (CLP PO)  or their authorized
       representative will conduct an on-site laboratory evaluation.  On-site
       laboratory evaluations are carried out to monitor the Contractor's
       ability to meet selected terms and conditions specified in the contract.
       The evaluation process incorporates two separate categories: Quality
       Assurance (QA) Evaluation and an Evidentiary Audit.

12.2   Quality Assurance On-Site Evaluation

       QA evaluators inspect the Contractor's facilities to verify the adequacy
       and maintenance of instrumentation, the continuity,  experience and
       education of personnel, and the acceptable performance of analytical and
       Quality Control  (QC) procedures.

12.2.1    The Contractor should expect that  items  to be  monitored  will  include,
          but not  be limited to,  the following:

          •    Size and appearance of  the facility;

          •    Quantity, age, availability,  scheduled maintenance, and
               performance of instrumentation;

          •    Availability, appropriateness, and utilization of the Quality
               Assurance Plan  (QAP)  and Standard Operating Procedures  (SOPs);

          •    Staff qualifications, experience, and personnel training
               programs;

          •    Reagents, standards,  and sample storage facilities;

          •    Standard preparation  logbooks and raw data;

          •    Bench sheets  and analytical logbook maintenance and review;  and

          •    Review of the Contractor's sample analysis/data package
               inspection/data  management procedures.

 12.2.2    Prior to an on-site evaluation,  various documentation pertaining to
          performance of the specific Contractor is integrated in a profile
          package for discussion during the evaluation.   Items that may be
          included are previous on-site reports,  performance evaluation sample
          scores,  Regional review of data,  Regional QA materials,  data audit
          reports, results of Contract Compliance Screening (CCS), and. data
           trend reports.

 12.3   Evidentiary Audit

        Evidence  auditors  conduct an  on-site  laboratory evaluation  to determine
        if laboratory policies and procedures are  in place to satisfy evidence
        handling  requirements as stated in Exhibit  F.  The evidence audit
        comprises a procedural audit, -an audit of  written Standard  Operating
        Procedures  (SOPs),  and an audit of analytical project file
        documentation.

 12.3.1     Procedural Audit.  The procedural audit consists of review and
           examination of actual SOPs and accompanying documentation for the
           following laboratory operations:   sample receiving, sample storage,

 OLC03.2                                   E-30

-------
                                                                  Exhibit E  -- Section  12
                                                   On-Site Laboratory Evaluations  (Con't)

          sample identification, sample security, sample tracking  (from receipt
          to completion of analysis), analytical project file organization and
          assembly, and proper disposal of samples and cogenerated wastes.

12.3.2    Written SOPs Audit.  The written SOPs audit consists of review and
          examination of the written SOPs to determine if they are accurate  and
          complete  for the following'laboratory operations: sample receiving,
          sample storage, sample identification, sample security, sample
          tracking  (from receipt to completion of analysis), and analytical
          project file organization and assembly.

12.3.3    Analytical Project File Evidence Audit.  The analytical project file
          evidence  audit consists of review and examination of the analytical
          project file documentation.  The auditors review the files to
          determine:

          •    The accuracy of the document inventory;

          •    The completeness of the file;

          •    The adequacy and accuracy of the document numbering system;

          •    Traceability of sample activity;

          •    Identification of activity recorded on the documents;  and

          •    Error correction methods.

12.4  Discussion of the On-Site Team's Findings

      "•5.The QA and evidentiary auditors discuss their findings with the USEPA
       Regional CLP PO prior to debriefing the Contractor.   During the
       debriefing,  the auditors present their findings and recommendations for
      ^corrective actions necessary to the Contractor personnel.

12.5  '-'-Corrective Action Reports for Follow-Through to Quality Assurance and
       Evidentiary Audit Reports

       Following an on-site laboratory evaluation,  QA and/or evidentiary audit
       reports which discuss deficiencies found during the on-site evaluation
       may be sent to the Contractor.   In a detailed letter,  the Contractor
       shall discuss the corrective actions implemented to resolve the
       deficiencies discussed during the on-site evaluation and discussed in
       the report(s) to the USEPA Regional CLP PO within 14 days of receipt of
       the report.

12.5.1    An alternate delivery schedule may be proposed by the Contractor,  but
          it is the sole decision of USEPA to  approve or disapprove the
          alternate delivery schedule.  If an  alternate delivery schedule is
          proposed, the Contractor  shall describe in a letter to the USEPA
          Regional  CLP PO and the Contracting  Officer why it is unable to meet
          the  delivery schedule listed in this section.  The USEPA Regional  CLP
          PO may grant an extension of up to 14 days for the Contractor's
          response  letter to the QA and evidentiary audit report.  An extension
          beyond 14 days must be approved by the USEPA Contracting Officer.
          The  Contractor shall proceed and not assume that an extension will be
          granted until so notified by the appropriate USEPA official.

12.5.2    If new SOPs are required to be written, or if existing SOPs are
          required  to be rewritten or amended  because of the deficiencies and
          the  subsequent corrective action implemented by the Contractor, the


                                           E-31                                   OLC03.2

-------
Exhibit E -- Section 12
On-Site Laboratory Evaluations  (Con't)

          Contractor shall write/amend the SOPs per the requirements listed in
          Section 6.

12.6   Corrective Actions

       If the Contractor fails to adhere to the requirements listed in this
       section, the Contractor may expect, but USEPA is not limited to, the
       following actions: reduction in the number of samples sent under the
       contract, suspension of sample shipment to the Contractor, an on-site
       laboratory evaluation, data package audit,  electronic data audit, a
       remedial performance evaluation sample, and/or contract sanctions.
 OLC03.2                                    E-32

-------
                                                                  Exhibit E -- Section 13
                                                                   Electronic Data Audits
13 . 0   ELECTRONIC DATA AUDITS

13 .1   Overview

       Periodically,  USEPA requests the instrument electronic data from
       Contractors for a specific Case in order to accomplish electronic data
       audits.   Generally, electronic data submissions and audits are requested
       for the following reasons.

       •   Program overview;

       •   Indication of  data  quality problems;

       •   Support for on-site audits;  and

       •   Specific  Regional requests.

13.1.1    Depending  upon  the reason  for an audit,  the  instrument  electronic
          data  from  a recent Case, a specific  Case, or a  laboratory  evaluation
          sample  may be requested.   Electronic data audits provide a mechanism
          to assess  adherence  to contractual requirements and to  ensure  the
          consistency of  data  reported on the  hardcopy/electronic deliverables
          with  that  generated  on analytical instruments.  This  function
          provides external monitoring of Program Quality Control (QC)
          requirements and checks adherence of the Contractor to  internal
          Quality Assurance  (QA)  procedures.   In  addition, electronic  data
          audits  enable USEPA  to evaluate the  utility,  precision, and  accuracy
          of the  analytical methods.

13.1.2    The Contractor  shall store all raw and  processed electronic
          analytical data in the appropriate instrument manufacturer's format,
          uncompressed, and with no  security codes.  The  data shall  include  all
          necessary  data  files for a complete  reconstruction of the  previously
          submitted  hardcopy and electronic deliverable data package.  All
          associated raw  data  files  in the instrument  manufacturer proprietary
          software format must be submitted if those files contain data  or
          instrumental parameters regarding any analysis  and or correction
          applied to an instrument or analytical  result.  This  instrument
          electronic data shall include data for  all samples and  all QC
          samples, including but not limited to:  blanks,  Matrix Spike/Matrix
          Spike Duplicate(s)  (MS/MSD),  Laboratory Control Sample  (LCS),
          instrument performance checks [4-Bromofluorobenzene  (BFB)  and
          decafluorotriphenylphosphine (DFTPP)],  initial  calibrations,
          Continuing Calibration,  as well as all  Contractor-generated  spectral
          libraries  and quantitation reports required  to  generate the  data
          package.   In addition,  the Contractor shall  supply raw  data  for the
          Method  Detection Limit (MDL)  studies and values for the year in which
          the Sample Delivery  Group  (SDG) was  analyzed.   The Contractor  shall
          maintain a written reference logbook of data files of EPA  sample
          number, calibration  data,  standards, blanks,  spikes,  and duplicates.
          The logbook shall  include  EPA sample numbers, identified by  Case and
          SDG.

13.1.3    Trie Contractor  is  required to retain the instrument electronic data
          for three  years after submission of  the reconciled Complete  SDG File.
          Electronic media shipped to USEPA designated recipient  must  be fully
          usable  by  the recipient.   Diskettes  must be  3.5 inch, high density,
          1.44  MB MS DOS  formatted and tapes must be either  4 mm  or  8  mm.
          Alternative means  for delivery of electronic data  may be utilized  by
          the Contractor  upon  prior  written approval by USEPA.  When submitting
          electronic instrument data to a USEPA,  the following  materials shall
          be delivered in response to the request.

                                           E-33
                                                                                  OLC03.2

-------
Exhibit E -- Section 13
Electronic Data Audits (Con't)


13.1.3.1      All  associated raw data files for all  analytical samples and all
              QC samples.   For example,  files for LCS,  blanks, initial and
              continuing  calibration standards and instrument performance check
              solutions (BFB and DFTPP).          ^

13.1.3.2      All  processed data files and quantitation output files associated
              with the raw data files described in Section 13.1.3.1.

13.1.3.3      All  associated identification and calculation files used to
              generate the data submitted in the data package.  This includes,
              but  is not  limited to, result files, acquisition files,
              calibration files, and method files.

13.1.3.4      All  Contractor-generated Mass Spectral library files  (NIST/EPA/NIH
              and/or Wiley, or equivalent, library not required).

13.1.3.5      A copy of the Contractor's written reference logbook relating data
              files to EPA sample number, LCS, BFB and DFTPP, calibration data,
              standards,  blanks, and spikes.  The logbook shall include EPA
              sample numbers and laboratory file identifiers for all samples,
              blanks, and standards, identified by Case and SDG.

13.1.3.6      A printout of the directory of all files in each directory,
              including all subdirectories and the files contained therein.

13.1.3.7      A copy  (hardcopy) of the completed Sample Data Package.

13.1.3.8      A statement attesting to the completeness of the electronic
              instrument data submission, signed and dated by the Contractor's
              laboratory manager.  The Contractor shall also provide a statement
              .attesting that the.data reported have not been altered in any way..
              These statements  shall be part of- a Cover Sheet that  includes the
              following information relevant to the data submission:

              •    Contractor name;

              •    Date of submission;

              •    Case number;

              •    SDG number;

              •    Instrument make and model number for each instrument;

              •    Instrument operating software name and version number;

              •    Data software name and version used for acquisition,
                   requantitation, and hardcopy/report generation;

              •    Data system computer;

              •    System operating software;

              •    Data system network;

              •    Data backup software;

              •    Data backup hardware;
 OLC03.2                                   E-34

-------
                                                                  Exhibit E  -- Section 13
                                                           Electronic Data Audits  (Con't)

              •   Media type and volume of data  (in MB) backed up,- and

              •   Names and telephone numbers of two Contractor contacts for
                  further  information regarding the submission.

13.2   Submission of the Instrument Electronic Data

       Upon request of the USEPA Regional Contract Laboratory Program (CLP)
       Project Officer (CLP PO),  the Contractor shall send the required
       instrument electronic data and all necessary documentation to USEPA
       designated recipient [e.g.,  Quality Assurance Technical Support (QATS)]
       within 7 days of notification.  An alternate delivery schedule may be
       proposed by the Contractor,  but it is the sole decision of USEPA to
       approve or disapprove the alternate delivery schedule.  If an alternate
       delivery schedule is proposed, the Contractor shall describe in a letter
       to the USEPA Regional CLP PO and the Contracting Officer why it is
       unable to meet the delivery schedule listed in this section.   The USEPA
       Regional CLP PO may grant an extension of up to 7 days for submission of
       the instrument electronic data.  An extension beyond 7 days must be
       approved by the USEPA Contracting Officer (CO).   The Contractor shall
       proceed and not assume that an extension will be granted until so
       notified by the appropriate USEPA official.

       NOTE: The instrument electronic data shall be shipped according to the
       procedures in Exhibit F.

13.3   Responding to the Electronic Data Audit Report

-       After completion of the electronic data audit,  USEPA may send a copy of
-       the electronic data audit report to the Contractor or may discuss the
      "electronic data audit report at an on-site laboratory evaluation.  In a
••--     detailed letter to the USEPA Regional CLP PO, the Contractor shall
•3-     ?discuss the corrective actions implemented to resolve the deficiencies
-     'listed in the electronic data audit report within 14 days of receipt of
       the report.

13.3.1    An  alternate delivery  schedule  may be proposed by  the Contractor,  but
          it  is  the  sole decision  of USEPA to approve  or disapprove  the
          alternate  delivery schedule.   If an alternate delivery  schedule is
          proposed,  the Contractor shall  describe  in a letter to  the USEPA
          Regional CLP PO  and the  Contracting Officer  why  it is unable to meet
          the delivery schedule  listed  in Section  13.3.  The USEPA Regional  CLP
          PO  may grant an  extension of  up to 14 days for the Contractor's
          response letter  to the electronic  data report.   An extension beyond
          14  days must be  approved by. the USEPA Contracting  Officer.  The
          Contractor shall proceed and  not assume  that an  extension  will be
          granted until so notified by  the appropriate USEPA official.

13.3.2    If  new Standard  Operating Procedures  (SOPs)  are  required to be
          written or SOPs  are required  to be amended because of the
          deficiencies and the subsequent corrective action  implemented by the
          Contractor, the  Contractor shall write/amend and submit the SOPs per
          the requirements listed  in Section 6.

i3.4   Corrective Actions

       If the Contractor fails to adhere to the requirements listed in Section
       13, the Contractor may expect, but USEPA is not limited to, the
       following actions: reduction in the number of samples sent under the
       contract, suspension of sample shipment to the Contractor,  an on-site
       laboratory evaluation,  an electronic data audit, a data package audit, a
       remedial laboratory evaluation sample, and/or contract sanctions.

                                           E-35                                    OLC03.2

-------
Exhibit E -- Section 13
Electronic Data Audits  (Con't)

13.5   Maintenance of the Magnetic Tape Storage Device

13.5.1    The Contractor  shall certify that the tape head alignment on the
          magnetic tape storage device is in compliance with the ANSI standards
          for nine track  magnetic  tapes.  If the Contractor does not have
          documentation of alignment within the last 12 months, the Contractor
          must perform or have performed the manufacturer's documented head
          alignment procedure within 60 days of contract award.  This is
          generally performed with a "skew" tape, certified to be in
          conformance with ANSI standards.  The alignment must be performed by
          qualified personnel.  The tape head alignment must be performed at a
          minimum once every 12 months, or whenever there is evidence that the
          tape head may be out of  alignment.

13.5.2   The tape system,  including recording head, must be  in  conformance with
         the manufacturer's physical and electrical standards.  Alignment of
         the remaining components  of the tape system  such as the  retracting
         arms,  must  be performed at intervals not  to  exceed  24  months.   If the
         Contractor cannot demonstrate  that  the remaining components of  the
         tape system are in alignment,  then  the Contractor must perform  or have
         performed the manufacturer's recommended alignment  procedure.
 OLC03.2      '                             E-36

-------
                                                                  Exhibit E -- Section 14
                                                                          Data Management
14.0   DATA MANAGEMENT

14.1   Overview

14.1.1    Data management procedures are defined as procedures specifying the
          acquisition or entry, update, correction, deletion, storage, and
          security of computer readable data and files.  These procedures shall
          be in written form and contain a clear definition for all databases
          and files used to generate or resubmit deliverables.  Key areas of
          concern include: system organization  (including personnel and
          security), documentation operations, traceability, and Quality
          Control  (QC).

14.1.2    Data manually entered from hardcopy shall be subject to QC checks and
          the error rates estimated.  Systems should prevent entry of incorrect
          or out-of-range data and alert data entry personnel of errors.  In
          addition, data entry error rates shall be estimated and recorded on a
          monthly basis by re-entering a statistical sample of the data entered
          and calculating discrepancy rates by data element.

14.2   Documenting Data Changes

       The record of changes in the form of corrections and updates to data
       originally generated,  submitted,  and/or resubmitted shall be documented
       to allow traceability of updates.   Documentation shall include the
       following for each change.

       •    Justification or rationale for  the change.

       •    Date and  initials of the person making  the  change(s).  Data changes
            shall be  implemented and reviewed by a  person or group independent
            of the  source generating the deliverable.

       •    Documentation of changes shall  be retained  according to the schedule
            of the. original deliverable.

       •    Resubmitted diskettes or other  deliverables shall be re-inspected as
            a part  of  the laboratory's internal inspection process prior to
            resubmission.  The  entire deliverable,  not  just  the changes, shall
            be inspected.

       •    The Laboratory Manager  shall approve changes to  originally submitted
            deliverables.

       •    Documentation of data changes may be requested by laboratory
            auditors.

14.3   Lifecycle Management Procedures

       Lifecycle management procedures shall be applied to computer software
       systems developed by the Contractor to be used to generate and edit
       contract deliverables.   Such systems shall be thoroughly tested and
       documented prior to utilization.

14.3.1    A software  test and  acceptance plan including test requirements, test
          results and acceptance criteria  shall be developed, followed, and
          available  in written form.

14.3.2    System changes shall not be made directly to production systems
          generating  deliverables.  Changes shall  be made first to a
          development  system and tested prior to implementation.
                                           E-37
                                                                                  OLC03.2

-------
Exhibit E -- Section 14
Data Management  (Con't)

14.3.3    Each version of the production system will be given an  identification
          number, date of installation, and date of last operation and will be
          archived.

14.3.4    System and operations  documentation shall be developed  and maintained
          for each system.  Documentation  shall include a user's  manual  and an
          operations and maintenance manual.

14.3.5    This documentation shall be available for on-site  review and/or upon
          written request by the USEPA Regional Contract Laboratory Program .
           (CLP)  Project Officer  (CLP PO).

14.4   Personnel Responsibilities

       Individual(s)  responsible for the following functions shall be
       identified.

       •    System operation and  maintenance including  documentation and
            training.

       •    Database  integrity,  including data  entry, data  updating and QC.

       •    Data and  system security,  backup and archiving.
 OLC03.2                                   E-38

-------
                EXHIBIT F
   CHAIN-OF-CUSTODY,  DOCUMENT  CONTROL
AND WRITTEN STANDARD OPERATING PROCEDURES
                   F-l                                    OLC03.2

-------
                           THIS  PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                    F-2

-------
                   Exhibit F  - Chain-of-iCustody, Document Control and
                          Written Standard Operating Procedures

Table of Contents



1.0    INTRODUCTION  	 5
       1.1    Purpose of Evidence Requirements  	•	5

2.0    STANDARD OPERATING PROCEDURES  	 6
       2.1    Sample Receiving   	6
       2.2    Sample Identification  	 7
       2.3    Sample Security	7
       2.4    Sample Storage   	 7
       2.5    Sample Tracking and Document Control  	 7
       2.6    Computer-Resident Sample Data Control 	 8
       2.7    Complete SDG File  (CSF) Organization and Assembly  	 9

3.0    WRITTEN STANDARD OPERATING PROCEDURES   	  11
       3.1    Sample Receiving   	  11
       3.2    Sample Identification  	  12
       3.3    Sample Security	  13
       3.4    Sample Storage	  13
       3.5    Sample Tracking and Document Control	  .  13
       3.6    Computer-Resident Sample Data Control 	  14
       3.7    CSF Organization and Assembly	15
                                           F-3                                    OLC03.2

-------
                            THIS PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                     p-4

-------
                                                                   Exhibit F -- Section 1
                                                                             Introduction
1.0    INTRODUCTION
       A sample is physical evidence collected from a facility or the
       environment.  Controlling evidence is an essential part of the hazardous
       waste investigation effort.  To ensure that U.S.  Environmental
       Protection Agency's (USEPA's) sample data and records supporting sample-
       related activities are admissible and have weight as evidence in future
       litigation. Contractors are required to maintain USEPA samples under
       Chain-of-Custody and to account for all samples and supporting records
       of sample handling, preparation, and analysis.  Contractors shall
       maintain sample identity, sample custody, and all sample-related records
       according to the requirements in this exhibit.

1.1    Purpose of Evidence Requirements

       The purpose of the evidence requirements include:

       •    Ensuring  traceability of samples while in possession  of  the
            Contractor;

       •    Ensuring  custody  of  samples while in possession of  the Contractor;

       •    Ensuring  the  integrity of sample identity while in  possession  of  the
            Contractor;

       •    Ensuring  sample-related activities are recorded on  documents or in
            other formats for USEPA sample receipt,  storage, preparation,
            analysis,  and disposal;

       •    Ensuring  all  laboratory records for each specified  Sample Delivery
            Group (SDG) will  be  accounted for when the project  is completed;  and

       •    Ensuring  that all laboratory records directly related to USEPA
            samples are assembled and delivered to USEPA or, prior to delivery,
            are available upon USEPA's request.
                                           F-5                                    OLC03.2

-------
Exhibit F -- Section 2
Standard Operating Procedures

2.0    STANDARD OPERATING PROCEDURES

       The Contractor shall implement the following Standard Operating
       Procedures (SOPs) for sample receiving, sample identification, sample
       security, sample storage, sample tracking and document control,
       computer-resident sample data control, and Complete Sample Delivery
       Group (SDG) File (CSF) organization and assembly to ensure
       accountability of USEPA sample Chain-of-Custody, as well as control of
       all USEPA sample-related records.

2.1    Sample Receiving

2.1.1     The  Contractor  shall  designate  a  sample  custodian responsible  for
          receiving  USEPA samples.

2.1.2     The  Contractor  shall  designate  a  representative to receive USEPA
          samples in the  event  that the sample custodian is not available.

2.1.3     Upon receipt, the condition of  shipping  containers and sample
          containers shall be inspected and recorded on Form DC-1 by the sample
          custodian  or a  designated representative.

2.1.4     Upon receipt, the condition of  the custody seals (intact/broken)
          shall be inspected and recorded on Form  DC-1 by the sample custodian
          or a designated representative.

2.1.5     The  sample custodian or a designated representative shall verify and
          record on  Form  DC-1,  the agreement or disagreement of information
          recorded on all documents received with  samples and information
          recorded on sample containers.

2.1.6     The sample custodian or a designated representative shall verify and •
          record the following information  on Form DC-1 as samples are received
          and inspected:

           •   Presence or  absence and condition of custody seals  on  shipping
               and/or sample containers;

           •   Custody  seal numbers  when present;

           •   Condition of the  sample bottles;

           •   Presence or  absence of airbills  or airbill  stickers;

           •   Airbill  or airbill sticker  numbers;

           •   Presence or  absence of Chain-of-Custody records;

           •   Sample tags  listed/not listed on Chain-of-Custody records;

           •   Presence or  absence of Traffic Reports  (TRs)  or Packing Lists;

           •   Presence or  absence of cooler temperature indicator bottle;

           •   Cooler temperature;

           •   Date of  receipt;

           •   Time of  receipt;

           •   EPA sample  numbers;
 OLC03.2                                    F-6

-------
                                                                   Exhibit F  -- Section 2
                                                    Standard Operating Procedures  (Con't)

          •   Presence or absence of sample tags;

          •   Sample tag numbers;

          •   Assigned laboratory numbers;

          •   Samples delivered by hand; and

          •   Problems and discrepancies.

2.1.7     The sample custodian or a designated representative  shall  sign, date,
          and record the  time  on all accompanying forms, when  applicable, at
          the time  of sample receipt (e.g.,  Chain-of-Custody records, TRs or
          packing lists,  and airbills).

          NOTE:  Initials  are not acceptable.

2.1.8     The Contractor  shall contact  the  Sample Management Office  (SMO) to
          resolve problems  and discrepancies including, but not  limited  to:
          absent documents; conflicting information;  absent or broken custody
          seals; and unsatisfactory sample  condition  (e.g., leaking  sample
          container).

2.-1.9     The Contractor  shall record resolution of problems and discrepancies
          by SMO.

2.2''    Sample Identification

2.2'-. 1  ;;   The Contractor  shall maintain the identity  of USEPA  samples and
       ~   prepared  samples  (including extracted samples, digested samples, and
          distilled samples) throughout the laboratory.
   •;;••  •'•»•••
2.2'-"2  ~:v  Each  sample and sample preparation container shall be  labeled  with
          the EPA sample  number or a unique laboratory sample  identification
          number.

2.3-    .Sample Security

2.3.1     The Contractor  shall demonstrate  that USEPA sample custody is
          maintained from receiving through retention or disposal.   A sample  is
          in custody if:

          •   It is in your possession;  or

          •   It is in your view after being in your possession;  or

          •   It is locked in a secure area after being in your possession;  or

          •   It is in a designated  secure area, accessible only to authorized
              personnel.

2.3.2     The Contractor  shall demonstrate  security of designated secure areas.

2.4    Sample Storage

       The Contractor shall designate storage  areas for USEPA  samples and
       prepared  samples.

2.5    Sample Tracking and Document  Control

2.5.1     The Contractor  shall record all activities  performed on USEPA
          samples.


                                            F-7                                    OLC03.2

-------
Exhibit F -- Section 2
Standard Operating Procedures  (Con't)

2.5.2     Titles which  identify the  activities  recorded shall be printed on
          each page of  all  laboratory documents.   (Activities include, but are
          not limited to: sample receipt;  sample  storage;  sample preparation,
          and sample analysis.)   When a  document  is  a  record of analysis, the
          instrument type and  parameter  group  (i.e., GC/MS-VOA) shall be
          included in the title.

2.5.3     When columns  are  used to organize  information recorded on laboratory
          documents, the information recorded  in  the columns shall  be
          identified in a column heading.

2.5.4     Reviewers' signatures shall be identified  on laboratory documents
          when reviews  are  conducted.

2.5.5     The laboratory name  shall  be  identified on pre-printed laboratory
          documents.

2.5.6     Each laboratory document entry shall be dated with the month/day/year
           (e.g.,  01/01/2000)  and signed (or  initialed)  by the  individual(s)
          responsible  for performing the recorded activity at  the time  the
          activity is  recorded.

2.5.7     Notations  on laboratory documents  shall be recorded  in  ink.

2.5.8     Corrections  to  laboratory  data reporting forms and raw  data shall  be
          made by drawing  single lines  through the errors and  entering  the
          correct information.  Information shall not  be obliterated or .
          rendered unreadable.  Corrections  and additions to  information shall
          be signed (or initialed) and dated.

2.5.9     Unused portions  of laboratory documents shall be lined-out.

2.5.10    Pages  in bound and.unbound logbooks shall be sequentially numbered.

2.5.11    Instrument-specific run logs, shall be maintained to  enable the
           reconstruction of run sequences.

2.5.12    Logbook entries shall be in chronological order.

2.5.13    Logbook entries shall include only one SDG per page,  except in the
           event where^SDGs "share" Quality Control  (QC) samples (e.g.,
           instrument run logs and extraction logs).

2.5.14    Information  inserted  into laboratory documents shall be affixed
           permanently  in-place.  The individual responsible for inserting
           information  shall sign and date across the insert and logbook page at
           the time information  is inserted.

 2.5.15    The Contractor shall  document disposal or retention of USEPA samples,
           remaining portions  of  samples,  and prepared  samples.

 2.6    Computer-Resident  Sample Data Control

 2.6.1     Contractor personnel  responsible  for original data entry  shall be
           identified at the time of data  input.

 2.6.2     The Contractor shall  make changes to electronic data in a manner
           which ensures that  the original data entry  is preserved,   the editor
           is identified, and  the revision date is recorded.
 OLC03.2                                    F-8

-------
                                                                   Exhibit F -- Section 2
                                                    Standard Operating Procedures  (Con't)

2.6.3     The  Contractor shall routinely verify  the accuracy of manually
          entered data, electronically entered data, and data acquired from
          instruments.

2.6.4     The  Contractor shall routinely verify  documents produced by the
          electronic data collection  system to ensure accuracy of the
          information reported.

2.6.5     The  Contractor shall ensure that  the electronic data collection
          system is secure.

2.6.5.1       The electronic data collection system shall  be maintained in a
              secure location.

2.6.5.2       Access to the electronic data collection system functions shall be
              limited to authorized personnel through utilization of software
              security techniques (e.g.,  log-ons or restricted passwords).

2.6.5.3       Electronic data collection systems shall be  protected from the
              introduction of external programs or software (e.g.,  viruses).

2.6.6     The  Contractor shall designate archive storage areas for electronic
          data and  the  software  required to access the data.

2.6.7     The  Contractor shall designate an individual responsible for
 ~         maintaining archives of electronic data, including the software.

2;. 6 .8     The  Contractor shall maintain the archives of electronic data and
          necessary software in  a secure location that shall be accessible only
          to authorized personnel.

2.7    Complete SDG File (CSF)  Organization and Assembly

2.7.1     The  Contractor shall designate a  document control officer  responsible
 -•    "fi  for  the organization and  assembly of the CSF.

2.7.2  .   The  Contractor shall designate a  representative responsible for the
          organization  and  assembly of the  CSF in the event that the document
          control officer  is not available.

2.7.3     The  Contractor shall maintain documents relating to the CSF in a
          secure location.

2.7.4-    All  original  laboratory forms and copies of SDG-related logbook pages
          shall be  included in the  CSF.

2.7.5     Copies of laboratory documents in the  CSF shall be photocopied in  a
          manner to provide complete  and legible replicates.

2.7.6     Documents relevant to  each  SDG including, but not limited  to, the
          following shall be included in the CSF:

          •   logbook pages;                         •  records  of  failed or
          •   bench sheets;                             attempted  analysis;
          •   mass spectra;                          •  custody  records,-
          •   chromatograms;                         •  sample tracking  records;
          •   screening records;                      •  raw data summaries;
          •   preparation records;                    •  computer printouts;
          •   repreparation records;                 •  correspondence;
          •   analytical records;                    •  FAX originals;
          •   re-analysis records;                    •  library  search results; and
                                                     •  other.


                                            F-9                                     OLC03.2

-------
Exhibit F -- Section 2
Standard Operating Procedures  (Con't)
2.7.7     The  document control  officer or  a  designated  representative  shall
          ensure  that sample  tags  are encased in  clear  plastic bags before
          placing them in the CSF.

2.7.8     CSF  documents  shall be organized and assembled  on an SDG-specific
          basis.

2.7.9     Original documents  which include information  relating  to more than
          one  SDG (e.g.,  Chain-of-Custody  records,  TRs, calibration logs) shall
          be filed in the CSF of  the lowest  SDG number, and copies of  these
          originals shall be  placed in the other  CSF(s).   The document control
          officer or a designated representative  shall  record the following
          statement on the copies  in (indelible)  dark ink:

                                           COPY
                     ORIGINAL DOCUMENTS ARE  INCLUDED IN CSF 	
                                                                       Signature
                                                                        Date

 2.7.10     All CSFs shall be submitted with a completed Form DC-2.   All
           resubmitted CSFs shall be submitted with a new or revised Form DC-2.

 2.7.11     Each item in the CSF and resubmitted CSFs shall be inventoried and
           assembled in the order specified on Form DC-2.  Each page of the CSF
           shall be stamped with a sequential number.  Page number ranges shall
           be recorded in the columns provided on Form DC-2.  Intentional gaps
           in the page numbering sequence shall be recorded in the "Comments"
           section on Form DC-2.   When inserting new or inadvertently omitted
           documents, the Contractor shall identify them with unique accountable
           numbers.  The unique accountable numbers and the locations of the
           documents shall be recorded in the "Other Records" section on Form
           DC-2.

 2.7.12     Before shipping each CSF, the document control officer or a
           designated representative shall verify the agreement of information
           recorded on all documentation and ensure that the information is
           consistent and the CSF is complete.

 2.7.13     The document control officer or a designated representative shall
           document the shipment of deliverable packages including what was
           sent, to whom the packages were sent, the date, and the. carrier used.

 2.7.14     Shipments of deliverable packages, including re-submittals, shall be
           sealed with custody seals by the document control officer or a
           designated representative in a manner such that opening the packages
           would break the seals.

 2.7.15     Custody seals shall be signed and dated by the document control
           officer or a designated  representative when sealing deliverable
           packages.
 OLC03.2                                    F-10

-------
                                                                   Exhibit  F  -- Section 3
                                                    Written Standard Operating Procedures
3.0    WRITTEN STANDARD OPERATING PROCEDURES
       The Contractor shall develop and implement the following written
       Standard Operating Procedures  (SOPs) for sample receiving, sample
       identification, sample security, sample storage, sample tracking and
       document control, computer-resident sample data control, and Complete
       Sample Delivery Group (SDG) File (CSF) organization and assembly to
       ensure accountability for USEPA sample Chain-of-Custody and control of
       all USEPA sample-related records.

3.1    Sample Receiving

3.1.1     The  Contractor  shall  have written  SOPs  for  sample  receiving which
          accurately reflect the  procedures  used  by the  laboratory.

3.1.2     The  written SOPs for  sample receiving shall ensure that  the
          procedures listed below are in-use at the laboratory.

3.1.2.1       The condition of shipping containers and sample containers are
              inspected and recorded on Form DC-1 upon receipt by the sample
              custodian or a designated representative.

3.1.2.2       The condition of custody seals are inspected and recorded on Form
              DC-1 upon receipt by the sample custodian or a designated
              representative.

3.1.2.3       The' presence or absence of the following documents/items
              accompanying the sample shipment is verified and recorded on Form
       -:•-      DC-1 by the sample custodian or a designated representative:

      ?'•'•      •  Custody  seals;
      :iv
 -•"•   ?:'      •  Chain-of-Custody records;

 ••"-'•    sA      •  Traffic  Reports  (TRs)  or  Packing Lists,-

 ~v-    -      •  Airbills or airbill  stickers;

              •  Sample tags; and

              •  Cooler temperature indicator bottle.

3.1.2.4       The agreement or disagreement of information recorded on shipping
              documents with information recorded on sample containers is
              verified and recorded on Form DC-1 by the sample custodian or a
              designated representative.

3.1.2.5       The following information is recorded on Form DC-1 by the sample
              custodian or a designated representative as samples are received
              and inspected:

              •  Presence or absence  and condition  of custody seals on shipping
                 and/or sample  containers;

              •  Custody  seal numbers when present;

              •  Condition of the sample bottles;

              •  Presence or absence  of airbills  or airbill  stickers;

              •  Airbill  or airbill sticker  numbers;
                                           F-ll                                    OLC03.2

-------
Exhibit F -- Section 3
Written Standard Operating Procedures (Con't)

              •  Presence or absence of Chain-of-Custody records;

              •  Sample tags listed/not listed on Chain-of-Custody records;

              •  Presence or absence of TRs  or Packing Lists;

              •  Presence or absence of cooler temperature indicator bottle;

              •  Cooler temperature;

              •  Date of receipt;

              •  Time of receipt;

              •  EPA sample numbers;

              •  Presence or absence of sample tags;

              •  Sample tag numbers;

              •  Assigned laboratory numbers;       :

              •  Samples delivered by hand;  and

              •  Problems and  discrepancies.

 3.1.2.6       The sample custodian or a designated representative shall sign,
              date, and record the time on all accompanying forms, when
              applicable,  at the time of sample receipt (e.g., Chain-of-Custody
              records, TRs or packing lists, and airbills).

              NOTE: Initials are not acceptable.

 3.1.2.7       The Contractor shall contact the Sample Management Office  (SMO) to
              resolve problems and discrepancies, including, but not limited to:
              absent documents; conflicting  information; absent or broken
              custody seals; and unsatisfactory sample condition  (e.g., leaking
              sample container).

 3.1.2.8       The Contractor shall record resolution of problems and
              discrepancies by SMO.

 3.2   Sample  Identification

 3.2.1     The Contractor shall have written SOPs for sample identification
           which accurately reflect the procedures used by the laboratory.

 3.2.2     The written SOPs for sample identification shall ensure that the
           procedures listed below are in use at the laboratory.

 3.2.2.1       The identity of  USEPA samples  and prepared  samples  is maintained
               throughout the laboratory when:•

               •   The Contractor assigns unique laboratory sample identification
                  numbers,  thus the written SOPs shall include a description of
                  the procedure used to assign these numbers;

               •   The Contractor uses prefixes or suffixes in addition to
                  laboratory sample identification numbers,  thus the written SOPs
                  shall include their definitions;  and
 OLC03.2                                    F-12

-------
                                                                   Exhibit F -- Section 3
                                            Written Standard Operating Procedures  (Con't)

              •   The Contractor uses methods to uniquely identify
                 fractions/parameter groups and matrix type, thus the written
                 SOPs shall include a description of these methods.

3.2.2.2       Each sample and sample preparation container is labeled with the
              EPA sample number or a unique laboratory sample identification
              number.

3.3    Sample Security

3.3.1     The Contractor  shall have written SOPs for sample  security which
          accurately reflect the procedures used by the laboratory._

3.3.2     The written SOPs for sample security  shall include the items listed
          below.

3.3.2.1       Procedures which ensure the following:

              •   Sample custody is maintained;  and

              •   The security of designated secure areas is maintained.

3.3.2.2       A list of authorized personnel who have access to locked storage
              areas.

3.4    Sample Storage

3.4.1     The Contractor  shall have  written SOPs for sample  storage which
      • •'•   accurately reflect the procedures used by the laboratory.

3.4.2     The written SOPs for sample storage shall describe locations,
          contents, and identities  of all  storage areas for  USEPA  samples and
          prepared samples in  the  laboratory.
  • '•-"  •>.••-.
3.5 •  ^Sample Tracking and Document Control

3.K-1 —.   The Contractor  shall have  written SOPs for sample  tracking and
          document control which accurately reflect the procedures used  by  the
          laboratory.

3.5.2     The written SOPs for sample tracking  and document  control shall
          include  the  items  listed  below.

3.5.2.1       Examples of all laboratory documents used during sample receiving,
              sample storage, sample transfer,  sample analyses, CSF organization
              and assembly,  and sample retention or disposal.

3.5.2.2       Procedures which ensure the following:

              •  All activities performed  on USEPA samples are recorded;

              •  Titles which  identify  the activities recorded are printed  on
                 each page of all laboratory documents;

              •   Information recorded in columns is identified with column
                 headings;

              •  Reviewers'  signatures  are identified on laboratory documents;

              •  The laboratory name is included on pre-printed laboratory
                 documents;
                                           F-13                                   OLC03.2

-------
Exhibit F -- Section 3
Written Standard Operating Procedures  (Con't)

              •  Laboratory document entries are signed arid dated with the
                 month/day/year  (e.g., 01/01/2000);

              •  Entries on all  laboratory documents are recorded in  ink;

              •  Corrections and additions to laboratory documents are made by
                 drawing single  lines  through the errors, entering the correct
                 information, and  initialing and dating the new information;

              •  Unused portions of laboratory documents are  lined-out;

              •  Pages in bound  and unbound  logbooks are sequentially numbered;

              •  Instrument-specific run  logs are maintained  to enable the
                 reconstruction  of run sequences;

              •  Logbook entries are recorded in chronological order;

              •  Entries are recorded  for only one  SDG per page, except  in  the
                 event where SDGs  "share" Quality Control  (QC) samples  (e.g.,
                 instrument run  logs and  extraction logs);

              •  Information inserted  in  laboratory documents is affixed
                 permanently,  signed or initialed,  and dated  across  the  insert;
                 and

              •  The retention or  disposal  of USEPA samples,  remaining portions
                 of samples, and prepared samples is documented.

 3 .6    Computer-Resident Sample  Data Control

 3.6.1     The Contractor shall have  written SOPs  for  computer-resident  sample
           data control which accurately reflect  the procedures used by the
           laboratory.

 3.6.2     The written SOPs for computer-resident  sample  data control shall
           include  the items listed below.

 3.6.2.1       Procedures which ensure  the following:

              •  Contractor personnel  responsible  for  original data entry are
                  identified;

              •  Changes to electronic data are made  such that the  original data
                 entry is preserved,  the  editor  is  identified, and  the  revision
                 date is recorded;

               •  The accuracy of manually entered data,  electronically entered
                 data,  and data acquired from instruments is verified;

               •   Report documents produced by the electronic data collection
                  system are routinely verified to ensure the accuracy of the
                  information reported;

               •   Electronic data collection system security is maintained;

               •   Archives of electronic data and accompanying software are
                  maintained in a secure location;  and

               •   Off-site backup and storage of  electronic data  is  maintained.
 OLC03.2                                    F-14

-------
                                                                   Exhibit F -- Section 3
                                            Written Standard Operating Procedures  (Con'ti

3.6.2.2       Descriptions of archive storage areas for the electronic data and
              the software required, to access data archives.

3.6.2.3       A list of authorized personnel who have access to electronic data
              collection system functions and to archived data".

3.7    CSF Organization and Assembly

3.7.1     The Contractor  shall have  written SOPs  for  CSF organization and
          assembly  which  accurately  reflect the procedures used by the
          laboratory.

3.7.2     The written  SOPs  for CSF organization and assembly shall ensure that
          the procedures  listed below are  in-use  at the laboratory.

          •   Documents relating to the CSF are maintained in a secure location;

          •   All original laboratory forms and copies of SDG-related logbook
              pages are included in the CSF;

          •   Laboratory documents are photocopied in a manner to provide
              complete and legible replicates;

          •   All documents relevant to each SDG are included in the CSF;

          •   Sample tags are encased in clear plastic bags by the document
              control officer or a designated representative before being placed
              in the CSF;

          •   The CSF is organized and assembled on an SDG-specific basis;

      -.«/   •   Original documents which contain information relating to more than
      vtf:       one SDG are filed in the CSF of the lowest SDG and copies are
              referenced to originals in the event that an original document
              contains information relating to more than one SDG;

      '-.   •   Each CSF is submitted with a completed Form DC-2, and re-submitted
              CSFs are submitted with a new or revised Form DC-2;

          •   Each page of the CSF is stamped with a sequential number and the
              page number ranges are recorded in the columns provided on Form
              DC-2.  Intentional gaps in the page numbering sequence are
              recorded in the "Comments" section of Form DC-2.   Inserted
              documents are recorded in the "Other Records" section of Form
              DC-2;

          •   Consistency and completeness of the CSF are verified by the
              document control officer or a designated representative;

          •   Shipments of deliverable packages are documented by the document
              control officer or a designated representative;

          •   Deliverable packages are shipped by the document control officer
              or a designated representative using custody seals in a manner
              such that opening the packages would break the seals; and

          •   Custody seals are signed and dated by the document control officer
              or a designated representative before placing them on deliverable
              packages.
                                           F-15                                   OLC03.2

-------
    EXHIBIT G
GLOSSARY OF TERMS
        G-l                                     OLC03.2

-------
                            THIS PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                    G-2

-------
                                                           Exhibit G -- Glossary of Terms
ALIQUOT - A measured portion of a field sample, standard, or solution, taken
for sample preparation and/or analysis.

ANALYSIS DATE/TIME - The date and military time (24-hour clock) of the
injection of the sample, standard, or blank into the GC/MS or GC system.

BAR GRAPH SPECTRUM - A plot of the mass-to-charge ratio  (m/e) versus relative
intensity of the ion current.

BLANK - An analytical sample designed to assess specific sources of
contamination.  See individual definitions for types of blanks.

BREAKDOWN - A measure of the decomposition of certain analytes  (DDT and
Endrin) into by-products.

4-BROMOFLUOROBENZENE  (BFB)  - The compound chosen to establish mass spectral
instrument performance  for  volatile  (VOA) analyses.

CALIBRATION FACTOR  (CF)  - A measure  of the gas chromatographic  response of a
target analyte  to the mass  injected.

CASE  - A finite, usually predetermined number of samples collected over a
given time period from  a particular  site.  Case numbers  are  assigned  by the
Sample Management Office (SMO).  A Case consists of one  or more Sample
Delivery Groups (SDGs).

CONTAMINATION  - A component of a  sample or an extract that is not
representative  of the environmental  source of the  sample.  Contamination may
stem .from other samples, sampling equipment, while in transit,  from laboratory
reagents, laboratory environment, or analytical instruments.

CONTINUING CALIBRATION  - Analytical  standard run every  12 hours to verify the
initial calibration of  the  system.

CONTINUOUS LIQUID-LIQUID EXTRACTION  - Used herein  synonymously  with the terms
continuous extraction,  continuous liquid  extraction, and liquid extraction.
This  extraction technique  involves boiling the extraction solvent in  a  flask
and condensing the  solvent  above  the aqueous sample.  The condensed solvent
drips  through  the  sample,  extracting the  compounds of interest  from the
aqueous phase.

DATE  - MM/DD/YYYY  - where  MM  =  01 for  January,  02  for February, ... 12  for
December; DD  = 01  to  31; YYYY =  1998,  1999,  2000,  2001,  etc.

DAY -  Unless  otherwise  specified, day  shall, mean  calendar day.

DECAFLUOROTRIPHENYLPHOSPHINE  (DFTPP) -  Compound chosen  to establish mass
spectral  instrument performance for  semivolatile  analysis.

DEUTERATED MONITORING COMPOUNDS (DMCs)  -  Compounds added to  every calibration
standard, blank,  and  sample used to  evaluate the  efficiency  of  the
extraction/purge  and  trap  procedures,  and the  performance of the  gas
chromatograph/mass  spectrometer (GC/MS)  systems.   DMCs  are isotopically
labeled (deuterated)  analogs  of native target  compounds.  DMCs are not
expected to  be naturally detected in the  environmental  media.
                                            G-3                                    OLC03.2

-------
Exhibit G -- Glossary of Terms

EXTRACTABLE - A compound that can be partitioned into an organic solvent from
the sample matrix and is amenable to gas. chromatography.  Extractables include
semivolatile  (SVGA) and pesticide/Aroclor compounds.

EXTRACTED ION CURRENT PROFILE  (EICP) - A plot of ion abundance versus time  (or
scan number) for ion(s) of specified mass(es).

FIELD SAMPLE - A portion of material to  be analyzed that is contained in
single or multiple containers and identified by a unique sample number.

GAS CHROMATOGRAPH  (GC) - The instrument  used to separate analytes on a
stationary phase within a chromatographic column.  The analytes are volatized
directly from the sample  (VOA), or injected as extracts  (SVA and PEST).  In
VOA and SVA analysis, the compounds are  detected by a Mass Spectrometer.  In
PEST analysis, the compounds are detected by an Electron Capture Detector.

IN-HOUSE - At the Contractor's  facility.

INITIAL CALIBRATION - Analysis  of analytical standards for a series of
different specified concentrations; used to define the linearity and dynamic
range of the response of the mass spectrometer or electron capture detector to
the target compounds.

INTEGRATION SCAN RANGE - The scan number of the scan at the beginning of the
area of integration to the scan number at the end of the area of integration.
Performed in accordance with Exhibit D VOA and SVOA.

INTEGRATION TIME RANGE - The retention time at the beginning of the area of
integration to the retention time at the end of the area of integration.

INTERNAL STANDARDS - Compounds  added to  every standard, blank, sample (for
volatiles), sample extract  (for semivolatiles), including Laboratory Control
Sample, at a known concentration, prior  to analysis.  Internal standards are
used as the basis for quantitation of the target compounds.

INSTRUMENT BLANK - A blank designed to determine the level of contamination
associated with the analytical  instruments.

LABORATORY - Synonymous with Contractor  as used herein.

LABORATORY CONTROL SAMPLE  (LCS) - The LCS is an internal laboratory quality
control sample designed to assess  (on an SDG-by-SDG basis) the capability of
the contractor to perform the analytical method.

m/z - Mass to charge ratio, synonymous with "m/e".

MATRIX - The predominant material of which the sample to be analyzed is
composed. For the purpose of this SOW, the sample matrix is water.

MATRIX EFFECT - In general, the effect of a particular matrix  (water), on the
constituents with which it contacts.  Matrix effects may prevent efficient
purging/extraction of target analytes, and may affect DMC and surrogate
recoveries.   In addition, nontarget analytes may be extracted from the matrix
causing interferences.
OLC03.2                                    G-4

-------
                                                           Exhibit G -- Glossary of Terms

MATRIX SPIKE - Aliquot of the water sample fortified  (spiked) with known
quantities of specific compounds and subjected to the entire analytical
procedure in order to indicate the appropriateness of the method for the
matrix by measuring recovery.

MATRIX SPIKE DUPLICATE - A second aliquot of the same sample as the matrix
spike  (above) that is spiked in order to determine the precision of the
method.

METHOD BLANK - An analytical control consisting of all reagents, internal
standards, and surrogate standards  (or DMCs for VGA and SV), that is carried
throughout the entire analytical procedure.  The method blank is used to
define the level of laboratory, background, and reagent contamination.

NARRATIVE  (SDG Narrative) - Portion of the data package which includes
laboratory, contract, Case and sample number identification, and descriptive
documentation of any problems encountered in processing the  samples, along
with corrective action taken and problem resolution.  Complete SDG Narrative
specifications are included in Exhibit B.

PERCENT DIFFERENCE  (%D)  - As used in this SOW and elsewhere  to compare two
values, the percent difference indicates both the direction  and the magnitude
of  the comparison, i.e., the percent difference may be either negative,
positive,  or zero.

PERFORMANCE EVALUATION MIXTURE  - A  calibration solution of specific analytes
used to evaluate both recovery and  percent breakdown  as measures of
performance.

PERFORMANCE EVALUATION SAMPLE  - An  external quality control  sample prepared by
USEPA  and is designed to assess the capability of the Contractor to perform
the analytical method.

PRIMARY QUANTITATION ION - A contract  specified  ion used  to  quantitate a
target analyte.

PROTOCOL  - Describes the exact procedures to be  followed  with respect to
sample receipt and handling, analytical methods, data reporting and
deliverables, and  document control.  Used synonymously with  Statement of  Work
 (SOW) .

PURGE  AND TRAP  (DEVICE)  - Analytical technique  (device) used to isolate
volatile  (purgeable) organics  by  stripping  the compounds  from water by a
stream of inert  gas, trapping  the  compounds  on an adsorbent  such as a porous
polymer  trap, and  thermally  desorbing  the trapped compounds  onto the gas
chromatographic  column.

PURGEABLES -  Volatile  compounds.

RECONSTRUCTED ION CHROMATOGRAM (RIC)  - A mass  spectral  graphical
representation  of  the  separation achieved by a gas  chromatograph;  a plot  of
total  ion current  versus retention time.

RELATIVE PERCENT DIFFERENCE (RPD)  - As used in this  SOW and elsewhere  to
compare  two values,  the relative  percent  difference  is  based on the mean  of
the two  values.

                                            G-5                                    OLC03.2

-------
Exhibit G -- Glossary of Terms

RELATIVE RESPONSE FACTOR (RRF) - A measure of the relative mass spectral
response of an analyte compared to its internal standard.  RRFs are determined
by analysis of standards and are used in the calculation of concentrations of
analytes in samples.

RELATIVE RETENTION TIME (RRT) - The ratio of the retention time of a compound
to that of a standard (such as an internal standard).

RESOLUTION - Also termed separation or percent resolution, the separation
between peaks on a chromatogram, calculated by dividing the depth of the
valley between the peaks by the peak height of the smaller peak being
resolved, multiplied by 100.

RESOLUTION CHECK MIXTURE - A solution of specific analytes used to determine
resolution of adjacent peaks; used to assess instrumental performance.

RESPONSE - or Instrumental Response - A measurement of the output of the GC
detector (MS, or EC)  in which the intensity of the signal is proportionate to
the amount  (or concentration) detected.  Measured by peak area or peak height.

RETENTION TIME (RT) - The time a target analyte is retained on a GC column
before elution.  The identification of a target analyte is dependent on a .
target compound's retention time falling within the specified retention time
window established for that compound.  Retention time is dependent on the
nature of the column's stationary phase, column diameter, temperature, flow
rate, and other parameters.

SAMPLE DELIVERY GROUP (SDG) - A unit within a sample Case that is used to
identify a group of samples for delivery.  An SDG is defined by the following,
whichever is most frequent:

       •      Each Case of field samples received,  or

       •      Each 20 field samples [excluding Performance Evaluation (PE)
              samples]  within a Case,  or

       •      Each 7  calendar day period during which  field samples in a Case
              are received (said period beginning with the receipt of the first
              sample  in the SDG).

In addition, all samples and/or sample fractions assigned to an SDG must have
been scheduled under the same contractual turnaround time.  Preliminary
Results have no impact on defining the SDG.

SAMPLE MANAGEMENT OFFICE  (SMO)  - A contractor operated facility operated under
the Contract Laboratory Analytical Services Support (CLASS) contract, awarded
and administered by USEPA.

SAMPLE NUMBER  (EPA Sample Number) - A unique identification number designated
by USEPA to each sample.  The EPA sample number appears on the sample Traffic
Report  (TR) which documents information on that sample.

SECONDARY QUANTITATION ION - Contract specified ion(s) to be used in
quantitation of target analytes when interferences prevent the use of the
primary quantitation ion.
OLC03.2                                    G-6

-------
                                                           Exhibit G -- Glossary of Terms

SEMIVOLATILE (SV) COMPOUNDS - Compounds amenable to analysis by extraction of
the sample with an organic solvent.  Used synonymously with Base/Neutral/Acid
(BNA)  compounds.

STANDARD ANALYSIS - An analytical determination made with known quantities of
target compounds; used to determine response factors.

STORAGE BLANK - Reagent water  (two 40.0 mL aliquots) stored with volatile
samples in an SDG.  It is analyzed after all samples in that SDG have been
analyzed; and it is used to determine the level of contamination acquired
during storage.

SULFUR CLEANUP BLANK -. A modified method blank that is prepared only when some  .
of the samples in a batch are subjected to sulfur cleanup.  It is used to
determine the level of contamination associated with the sulfur cleanup
procedure.  When all of the samples are subjected to sulfur cleanup, then the
method blank serves this purpose.  When none of the samples are subjected to
sulfur cleanup, no sulfur cleanup blank is required.

SURROGATES  (Surrogate Standard)  - For pesticides/Aroclors, compounds added to
every blank, sample, including Laboratory Control Sample, requested MS/MSD,
and standard; used to evaluate analytical efficiency by measuring recovery.
Surrogates are not expected to be detected in environmental media.

TARGET COMPOUND  LIST  (TCL) - A list of compounds designated by the Statement
of Work  (Exhibit C) for analysis.

TENTATIVELY IDENTIFIED COMPOUNDS (TIC) - Compounds  detected in samples that
are not target compounds, internal standards, deuterated monitoring compounds,
or surrogates.   Up to 30 peaks,  not including those identified as alkanes
 (those greater than 10% of the peak area or height  of the nearest internal
standard),  are subjected to mass spectral library searches for tentative
identification.

TIME  - When required to record time on any deliverable item,  time shall be
expressed as Military Time,  i.e.,  a 24-hour clock  (0000-2359).

TRAFFIC REPORT  (TR) - A USEPA  sample  identification form  filled out by the
sampler,  which accompanies the sample during shipment to  the  laboratory and
which documents  sample condition and  receipt by the laboratory.

TWELVE-HOUR TIME PERIOD - The  twelve  (12)-hour time period for GC/MS system
instrument  performance check,  standards calibration (initial  or continuing
calibration),  and method blank analysis begins at the moment  of injection of
the DFTPP or BFB analysis that the laboratory submits as  documentation of
instrument  performance.  The time period  ends after 12 hours  have elapsed
according to the system clock.   For pesticide/Aroclor analyses performed by
GC/EC,  the  12-hour  time period in the analytical  sequence begins at the moment
of  injection of  the instrument blank  that precedes  sample analyses, and ends
after twelve hours  have elapsed  according to the  system clock.

VALIDATED TIME OF SAMPLE RECEIPT (VTSR)  - The date  on which a sample is
received at the  Contractor's facility,  as recorded  on the shipper's delivery
receipt  and Sample  Traffic Report.
                                            G-7                                    OLC03.2

-------
Exhibit G -- Glossary of Terms

VOLATILE COMPOUNDS - Compounds amenable to analysis by the purge and trap
technique.   Used synonymously with purgeable compounds.

WIDE BORE CAPILLARY COLUMN - A gas chromatographic column with an internal
diameter (ID) that is greater than or equal to 0.53 mm.   Columns with lesser
diameters are classified as narrow bore capillary columns.
OLC03.2                                    G-8

-------
           EXHIBIT H
AGENCY STANDARD IMPLEMENTATION
              H-l                                    OLC03.2

-------
                            THIS PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                     H-2

-------
                       Exhibit  H - Agency Standard Implementation

                                    Table of Contents


Section                                                                          .    Page

1.0    FORMAT CHARACTERISTICS	.  .  5

2.0    RECORD TYPES	7

3.0    PRODUCTION RUNS  .  .	8

4.0    RECORD SEQUENCE  	  9

5.0    FILE/RECORD INTEGRITY  	   10

6.0    DATES AND TIMES	10

7.0    MULTIPLE VOLUME DATA	   10

8.0    DELIVERABLE  	   11

9.0    RECORD LISTING	.  .	12
       9.1    Production Run Header Record  (Type 10)   	   12
       9.2    Chromatography Record (Type 11)	   13
       9.3    Sample Header Data Record  (Type 20)	   14
       9.4    Sample Header Data Record  (Type 21)	   16
       9.5    Sample Condition Record  (Type 22)  	   18
       9.6    Associated Injection and Counter Record  (Type 23)  	   19
       9.7    Sample Cleanup Record (Type 27)  	   22
       9.8    Results Data Record (Type  30)	   23
       9.9    Auxiliary Data Record (Type 32)	   26
       9.10   Name Record (Type' 33)	  '.  .   27
       9.11   Instrumental Data  Readout  Record  (Type 36)   	:  .  .  .  .   28
       9.12   Comment Record  (Type 90)   	   30

10.0   DEFINITIONS OF VARIOUS CODES USED IN AGENCY STANDARD RECORDS  	   31
       10.1   Quality Control and Related Codes  (QCC)  in Type 20 Records   	31
       10.2   Codes For Sample Medium  (Matrix, Sources)  	   33
       10.3   List of Sample and Result  Qualifiers	33
                                            H-3                                    OLC03.2

-------
                            THIS PAGE INTENTIONALLY LEFT BLANK
OLC03.2                                    H-4

-------
                                                                   Exhibit H -- Section 1
                                                                   Format Characteristics
1.0    FORMAT CHARACTERISTICS
1.1    This constitutes an implementation of the USEPA Agency Standard for
       Electronic Data Transmission based upon analytical results and ancillary
       information required by the contract.  All data generated by a single
       analysis are grouped together,  and the groups are aggregated to produce
       files that report data from a Sample Delivery Group (SDG).  Because this
       implementation is only a subset of the Agency Standard,  some fields have
       been replaced by delimiters as place holders for non-Contract Laboratory
       Program (non-CLP) data elements.

1.2    This implementation includes detailed specifications for the required
       format of each record.  The position in the record where each field is
       to be contained relevant to other fields is specified, as well as the
       maximum length of the field.  Each field's required contents are
       specified as literal  (contained in quotes), which must appear exactly as
       shown (without quotes), or as a variable for which format and/or
       descriptions are listed in the format/contents column.  Options and
       examples are listed for most fields.  For fields where more than three
       options are available, a list and description of options are supplied on
       a separate page following the record descriptions.  Fields are separated
       from each other by the delimiter "|"  (ASCII 124).  Fields that do not
       contain data should be zero length or a blank field (empty with no space
       or additional delimiters between the delimiters before and after the
       field) with the delimiter as a place holder.  For the purposes of
       Section 9 of this exhibit, wherever  "blank" is given as an option under
       the "Format/Contents" column, it refers to a blank field as explained
       above.

1.3    Numeric fields may contain numeric digits, a decimal place, and a
       leading minus sign.  A positive sign is assumed if no negative sign is
       entered in a numeric  field and shall not be entered into any numeric
       field.  Values that exceed the maximum length allowed shall be reported
       to the maximum possible, maintaining the specified decimal place and
       maximum field length  restrictions.

1.4    Requirements for significant figures and number of decimal places are
       specified in Exhibit  B.  The numeric field lengths are specified such
       that  all possible numeric values can be written to the file.  The size
       of the numeric field  indicates the maximum number of digits, including a
       decimal place and negative  sign  (if  appropriate), that can appear in the
       field at the same time.  Therefore,  the number reported may need to be
       rounded  (using rounding  rules described in Exhibit B) to  fit into the
       field.  The rounding  shall maintain  the greatest  significance possible
       providing the field length  limitation.  In addition,  the  rounded number
       that  appears on  the form, and therefore in the field  on the diskette
       file, must be used  in any calculation  that may result in  other numbers
       reported on the  same  form or other  forms  in  the  SDG.  The numbers/values
       reported by the  Contractor  are .used  by CCS to calculate a result  [e.g.,
       Contract Required Quantitation Limit (CRQL)].  The  final  value
       calculated .by CCS is  then rounded according  to rounding rules described
       in Exhibit B and is used for comparison to the final value  (e.g., CRQL)
       reported by the  Contractor.  Field  lengths should only be as long as
       necessary to contain  the data; packing with  blanks  is not allowed.

 1.5    USEPA is  currently  developing  a  data delivery  strategy that may be used
       as  an alternative to  the requirements  stated in  Exhibit H.  This
       strategy's  intent is  to  provide  a neutral  data delivery structure to the
       Contractor  that  will  further  facilitate  the  exchange  of analytical
       information generated under this analytical  protocol.  The proposed
       strategy  is  intended  to  accommodate  laboratories that generate data
       transmission  files  under multiple data formats.   Upon implementation of


                                            H-5                                    OLC03.2

-------
Exhibit H -- Section 1
Format Characteristics  (Con't)

       this alternate electronic data delivery strategy by the CLP and prior to
       submission of data in alternate format(s),  the Contractor must first
       demonstrate its ability to provide electronic data as stated in this
       Exhibit H, and obtain written permission from the CLP for the submission
       of data in alternate format(s).  The Contractor will receive a written
       response to its request within 90 calendar days. However, until the
       implementation of this alternate electronic data delivery strategy by
       the CLP, all electronic data deliverables must be provided as specified
       in this Exhibit H.
OLC03.2                                    H-6

-------
                                                                   Exhibit H -- Section 2
                                                                             Record Types
2.0    RECORD TYPES
2.1    The Agency Standard consists of variable length ASCII records.  Maximum
       field length specifications match the reporting requirements in Exhibit
       B.  The last two bytes of each record shall contain "carriage return"
       and "line feed", respectively.

2.2    This implementation consists of twelve record types that can be
       summarized in four groups, designated by the first record type in each
       group:

        Type                 Type  ID    Contents

        Run Header              10      Information pertinent to a group of
                                        samples processed in a continuous
                                        sequence; usually several per SDG
        Sample Header           20      Sample identifying, qualifying, and
                                        linking information
        Results Record   .       30      Analyte results and qualifications
        Comments Record         90      Free form comments

2.3    A separate run header is used for volatiles (VOA) , semivolatiles (SV),
       and for each column analysis for pesticides (PEST)  [minimum of four type
       10 series for VOA/SV/PEST Sample Delivery Group  (SDG)].   The 20 series
       records contain sample characteristics and link samples within an SDG to
       the corresponding calibrations, blanks, and other Quality Controls
       (QCs).  The 30 series records contain the actual analytical results by
       analyte within each sample.  The 10, 20, and 30 records are associated
       with each other by their position in the file  (i.e., 30 series records
       follow the corresponding 20 series, which in turn follow the 10 series
       run header records).
                                            H-7                                    OLC03.2

-------
Exhibit H -- Section 3
Production Runs

3 . 0    PRODUCTION RUNS

3.1    A production run represents a "group" or "batch" of samples that are
       processed in a continuous sequence under relatively stable conditions.
       Specifically:

3.1.1     Calibration  - All  samples  in a  run use  the  same  initial calibration
          data.

3.1.2     Method number -  Constant throughout  a run.

3.1.3     Instrument conditions  - Constant  throughout a run.

3.2    Each instrumental analysis consists of a separate production run and is
       reported in a separate file.   There will be a separate production run
       for each of the two pesticide Gas Chromatograph (GC) columns utilized.
       Thus, a full three fraction analysis will consist of a minimum of four
       production runs.

3.3    Example of the Sequence of Record Types in a File

    10                 Contains Run Header information.
        11            Contains additional run-wide information.
        20            Occurs once for each sample, calibration,  mean response
                      factor, matrix spike duplicate result,  etc.  Acts as a
                   •  header.
        21
      ,  22            Contains additional information for samples.
        23
      it •
     $'• 27
             30        Occurs once for each final analytical result.   Reports
                      the value being determined as defined by the type 20.
                 32   Reports any auxiliary data necessary.
                 33   Reports compound names for tentatively identified.
                      compounds  (TICs)  if necessary.
                 36   Reports any instrumental data necessary.
             30        Values for the next analyte or parameter being measured.
                 32   Additional data may vary for each parameter,  and may
                 33   occur in any order.   Multiple occurrences of the same
                 36   record type, however,  must be consecutive.
             30        Continues for  as many as are necessary.
                 32
                 33
                 36
             30
                 32
                 33
                 36
        20            Next Sample Header record.   The following applies to the
        21            next sample or other group of data.
        22
             30
                 32
                 33
                 36
             30
                 32
                 33
                 36
OLC03.2                                    H-8

-------
                                                              Exhibit H -- Sections 3 & 4
                                                                          Record Sequence
                     etc.
        20
        21
            30
                 32
                 33
              •   36
                      etc.
4.0    RECORD. SEQUENCE

4.1    The sequence of records for Agency Standard files is as follows:  A Run
       Header (type 10) record shall be present once and once only (per file)
     .  as the first record in a file.  Therefore, a complete VOA/SV/PEST SDG
       will consist of several files.

4.2   • Each environmental sample, calibration standard, or Quality Control (QC)
       sample is represented by a group composed of type 20, 21, 22,  23, and 27
       records,  that hold sample level identifying information, followed by
       type 30,  32, 33, and 36 records for each method analyte including
       surrogates, deuterated monitoring compounds, and internal standards in
       the sample.  The type 20 record holds a count for the number of method
       analytes being determined and includes all target compounds, surrogates,
       deuterated monitoring compounds, and internal standards plus each peak
       of the multi-component pesticides  [do not include Tentatively Identified
       Compounds  (TICs) in this count].  A separate field on the type 23 record
       contains the number of TICs found.  Type 20 records shall occur in the
       order of sample analysis.  In addition, a type 20 record with a QC code
       "MNC", followed by a type 30 record for each method analyte (reporting
       values such as mean response factors) will appear after the type 10 or
       type 11 record and before the type 20 record that initiates the
       analytical sequence.  Similarly, for pesticide runs, a type 20 record
       with a Quality Control  (QC) code "FLO" for Florisil recovery,  followed
       by type 30 records for each of  the method analytes  (and the two
       surrogates) included in the Florisil check will appear before the type
       20 record that  initiates the analytical sequence.

4.3    Type 90 comment records may be  defined to occupy any position after the
       type 10  (header) record.
                                            H-9                                    OLC03.2

-------
Exhibit H -- Sections 5-7
File/Record Integrity

5.0    FILE/RECORD INTEGRITY

       All record types shall contain the following check fields to ensure file
       and record integrity:

       Record       Field   Field
       Position     Length  Contents

       First Field      2    Record type                 "10" or as appropriate
       Last Field       5    Record sequence number      00001-99999,  numbered
                                                        within file sequentially
                        4    Record checksum1            Four hexadecimal digits
                        2    Must contain CR and LF  .

6.0    DATES AND TIMES

       Date or time-of-day information consists of successive groups of two
       decimal digits (except year, which is four decimal digits), each
       separated by delimiters.  Dates are given in the order YYYY MM DD, and
       times as HH MM.  All hours shall be given as 00 to 23 using a 24-hour
       clock and shall be local time.  All days shall be given as 01 to 31.
       All months shall be given as 01 to 12 (e.g., Ql is January, 02 is
       February).

7.0  _  MULTIPLE VOLUME DATA
     2"'. .

       There is no requirement under this format that all the data from an
     "  entire Sample Delivery Group  (SDG) fit onto a single diskette.  However,
       each single production run must fit onto a single diskette if possible.
     ~" If that is not possible, then it is necessary that all files start with
       a type 10 record, and that the multiple type 10 records for each file of
       the same production run be identical.  Information for a single sample
       shall not be split between files.
       1The checksum is the sum of the ASCII representation of the data on.the
record up to the Record Sequence Number  (not'including the Record Sequence
Number) plus the checksum of the previous record.  The sum is taken modulo
65536  (216)  and is  represented as four hexadecimal  digits  (i.e.,  the  remainder
of the sum divided by  65536 represented as  four hexadecimal digits).

OLC03.2                                    H-10

-------
                                                                   Exhibit H -- Section 8
                                                                              Deliverable
8. 0    DELIVERABLE
8.1    The file shall be submitted on IBM-compatible,  3.5 inch high density
       1.44 MB diskettes.  The diskettes shall be formatted and recorded using
       MS-DOS Operating System.  The diskettes shall contain all information
       relevant to one and only one Sample Delivery Group (SDG).  An
       alternative means of electronic transmission may be utilized if approved
       in advance by the USEPA.

8.2    Agency Standard data from an entire SDG may not fit onto a single
       diskette.  If a single production run is being split onto multiple
       diskettes, then all files shall start with a type 10 record, and the
       multiple type 10 records for each file of the same production run shall
       be identical.  Do not split the data from a single sample onto multiple
       diskettes.

8.3    Information on the diskette must correspond to information submitted in
       the hardcopy raw data package and on the hardcopy raw data package
       forms.  For example, type 30 results field specifies maximum length of
       13.  When reporting Contract Required Quantitation Limits  (CRQLs) or
       results on Form 1, maximum length is 13 as is specified in this exhibit;
       when reporting  'calculated amounts' on Form 7D, hardcopy specified
       maximum length is 8.  Unused records shall not be included.on the
       diskettes.  If the information submitted in the hardcopy data package
       forms is  changed, the information in the electronic file  (e.g.,
       diskette) shall be changed accordingly, and a complete electronic
       deliverable containing  all the information for the SDG shall be
       resubmitted along with  the hardcopy at no additional cost to the USEPA.

8.4    Each diskette shall be  identified with an external label containing  (in
       this order) the following information:

       •    Disk Density;
       •    File Name(s);
       •    Laboratory Name (optional);
       •    Laboratory Code;
       •    Contract Number;
       •    Case Number/SDG;
       •    Client Number (where applicable); and
       •    Initial Submission or Resubmission (as applicable)  and Date.

8.5    The format  for  File Name shall be XXXXX.001 to XXXXX.O99.  Where XXXXX
       is  the  SDG  identifier,  0 designates Organics, and 01 through 99  is the
       file number.

8.6    Dimensions  of the label must be  in the range of 2-1/2" to  2-3/4" long by
       2"  to  2-1/8" wide for a 3-1/2 inch IBM-compatible diskette.

8.7    Section 9.0  (Record Listing) provides  information for the  usage  of each
       of  the  record types.  Where  specified, labels indicate the nature of the
       value(s)  that follow on that record.   If  the value(s) will not be
       reported, the label shall be omitted.

8.7.1     A record type 30 for each Target Compound List (TCL)  compound,
           surrogate, deuterated monitoring compound,  and internal standard
           shall be reported.   If the TCL is not detected,  the 'U' qualifier in
           the appropriate field shall be indicative of that.

8.7.2     For multicomponent analytes  (Aroclors/toxaphene),  if the
           multicomponent analyte is detected, a record type 30 and 32 shall be
           reported for each peak identified.
                                           H-ll                                    OLC03.2

-------
Exhibit H -- Section 9
Record Listing

9.0    RECORD LISTING

       The following lists every record type required to report data from a
       single Sample Delivery Group (SDG).

9.1    Production Run Header Record (Type 10)

       Use:       Each production  run  will  start  with a  record type 10.
                    CONTENTS                          FORMAT/CONTENTS
                    RECORD TYPE           .            "10"
                    Delimiters                      .  ||||||
                    INSTRUMENT/DETECTOR               Character2
                    Delimiter                         |
                    METHOD NUMBER                   .  Character3
                    Delimiters                        ||
                    LAB CODE         •                 Character
                    Delimiters                        | | |  |
                    CONTRACT NUMBER                   Character
                    Delimiter                         |
                    INSTRUMENT ID                     Character
                    Delimiters                        ||
                    LABORATORY NAME                   Character
                    Delimiters                        ||
                    RECORD SEQUENCE NUMBER            Numeric
                    CHECKSUM                          Character
       2General descriptor (GC/MS for VOA/SV analysis or GC for pesticide
analysis  on GC/ECD).

       3OLC03.1V for volatiles; OLC03.1B for semivolatiles; OLC03.1P for
pesticides.   (0  for Organic, L  for Low, C  for Concentration,  zero three point
zero  for  document number, V  for volatiles, B for  semivolatiles,  P for
pesticides.)

OLC03.2                                    H-12

-------
                                                                   Exhibit H  -- Section 9
                                                                   Record Listing  (Con't)
9.2    Chromatography Record (Type 11)

       Use:        To describe  chromatograph  condition.   Must  be present  once for
                   each  production run immediately following the record type 10.
                    CONTENTS                          FORMAT/CONTENTS
                    RECORD TYPE                       "11"
                    Delimiter                         |
                    GC COLUMN IDENTIFICATION          Character
                    Delimiters                        | |
                    GC COLUMN ID4                      Numeric  (mm)
                    Delimiters   .                     I I  I I I I  I II II
                    RECORD SEQUENCE NO.                Numeric
                    CHECKSUM                          Character
        * Internal Diameter  of  the GC column used.

                                           H-13                                    OLC03.2

-------
Exhibit H -- Section 9
Record Listing  (don't)

9.3    Sample Header Data Record  (Type 20)
        MAXIMUM
        LENGTH	
               2
               2
              12

               1
               1
               1
               3
               1
               3
               1
               5
               1
               6
               1
               4

               1
               2

               1
               2
               1
               2

               1
               2

               2
               2
               1
               5
CONTENTS
RECORD TYPE
Delimiters
EPA SAMPLE NUMBER

Delimiter
MATRIX
Delimiter
QC CODE
Delimiter
SAMPLE QUALIFIER
Delimiter
CASE NUMBER
Delimiter
SDG NO.
Delimiter
SAMPLE/BLANK/STANDARDS YEAR
ANALYZED
Delimiter
SAMPLE/BLANK/STANDARDS MONTH
ANALYZED
Delimiter
SAMPLE/BLANK/STANDARDS DAY ANALYZED
Delimiter
SAMPLE/BLANK/STANDARDS HOUR
ANALYZED
Delimiter
SAMPLE/BLANK/STANDARDS MINUTE
ANALYZED
Delimiters
SAMPLE VOL UNITS
Delimiter
SAMPLE VOL
FORMAT/CONTENTS
"20"
II
As is exactly on the
hardcopy form
I
CHARACTER5
I
Character (See Section 10)
I
RIN/REX/REJ/SRN/blank6
I
Numeric
I
Character
I
YYYY
MM

I
DD
I
HH

I
MM

II
"ML"/blank7
I
Numeric8
       5"0" if not applicable  (calibration, tune, etc.); "1" for water.

       6"RIN" for reinjection,- "REX" for re-extractions; "REJ" for rejected
 samples,-  "SRN"  for dilutions;  and leave blank (empty  field  with  zero length)
 when none  of  the  previous conditions apply.   In  case  of multiple operations  on
 a  sample,  the final  operation will be indicated  (e.g.,  reinjection of a
 dilution;  AAA12DLRE  would have a QC Code of  "RIN").

       7Sample VOL unit is ML  (milliliters) for liquids.  Leave blank  (zero
 length)  if not  applicable.

       8Sample VOL is the volume  in milliliters for liquid.   Sample VOL
 includes the  purge volume.
OLC03.2
                                           H-14

-------
       Sample Header Data Record  (Type 20)  (Con't)
       MAXIMUM
       LENGTH-
              1
              3
              3
              5
              4
CONTENTS
Delimiter
ANALYTE COUNT
Delimiters
RECORD SEQUENCE NO.
CHECKSUM
                                                                  Exhibit H  --  Section 9
                                                                  Record Listing  (Con't)
FORMAT/CONTENTS
I
Numeric9
III
Numeric
Character
       9Counts  TCL analytes,  surrogates,  deuterated monitoring  compounds  (DMC),
internal standards, and all peaks reported for multi-component PCBs.  Do not
include the count for TICs in this field.  For calibrations, also count DFTPP,
if included in calibration solution.

                                           H-15
                                                                                  OLC03.2

-------
Exhibit H -- Section 9
Record Listing  (Con't)

9.4    Sample Header Data Record  (Type 21)

       Use:        Continuation of Type 20.
       Position:  Follows  the  Type 20 to  which  it  applies.
        MAXIMUM
        LENGTH     CONTENTS
           2       RECORD TYPE
           1       Delimiter
           1       PURGE
           3       Delimiters
           1       EXTRACTION
           2       Delimiters
           6       CLIENT NUMBER
           1       Delimiter
           14       LAB  FILE/SAMPLE  ID
           1       Delimiter
           4       YEAR EXTRACTED
           1       Delimiter
           2       MONTH EXTRACTED
           1       Delimiter
           2       DAY  EXTRACTED
           2       Delimiters
           4       YEAR RECEIVED

           1       Delimiter
           2       MONTH RECEIVED

           1       Delimiter
           2       DAY  RECEIVED

           1       Delimiter
           7       IDENTIFICATION/LOT
                   NUMBER OF LCS
           1       Delimiter
FORMAT/CONTENTS
"21"
I
"N" (for VOA); blank (for SV or PEST10
III
S/C/H/blank  (for all volatile samples)11
II
Character
I
Character12
I
YYYY/blank (for volatiles)
I
MM/blank  (for volatiles)
I
DD/blank  (for volatiles)
II
YYYY/blank (for standards, tunes, and
blanks)
I
MM/blank  (for standards, tunes, and
blanks)
I
DD/blank  (for standards, tunes, and
blanks)
I
Character13
       10"N"  for not heated purge.   All low concentration volatile samples are
required to be  purged  at  ambient temperature.

       11"S"  for separatory funnel;  "C" for continuous liq-liq without
hydrophobic membrane;  "H"  for  continuous  liq-liq with hydrophobic membrane,-
blank  (zero length  field)  for  all volatile samples.

       12Lab  File ID for volatile and semivolatile  analyses.   Lab Sample ID for
pesticides in same  format  as on forms.

       13lf the  LCS  solution was purchased by the Contractor from a third
party, report the identification number used by the  third party to identify
the LCS lot, if available  (Form 3).   If the LCS solution was  prepared in-
house, leave this entry blank.
OLC03.2
                                           H-16

-------
                                                            Exhibit  H --  Section 9
                                                            Record Listing (Con't)
Sample Header Data Record  (Type 21)  (Con't)
MAXIMUM
LENGTH      CONTENTS                      FORMAT/CONTENTS
    8       INJECTION VOLUME              Numeric/blank (for VGA)"
    2       Delimiters                    ||
    5       RECORD SEQUENCE NO.            Numeric
    4       CHECKSUM                      Character
14Injection volume,  in  uL/  for SVs and  PESTs.

                                    H-17                                   OLC03.2

-------
Exhibit H -- Section 9
Record Listing  (Con't)

9.5    Sample Condition Record  (Type 22}

       Use:         Continuation of type 20.  Used to describe additional Sample
                   Conditions.
       Position:   Follows the  type 20  and 21 to .which it  applies.
                    CONTENTS
                    RECORD TYPE
                    Delimiter
                    CALIBRATION YEAR
                    Delimiter
                    CALIBRATION MONTH
                    Delimiter
                    CALIBRATION DAY
                    Delimiter
                    CALIBRATION HOUR
                    Delimiter
                    CALIBRATION MINUTE
                    Delimiter
                    CALIBRATION FILE ID
                    Delimiters
                    EXTRACT VOLUME
                    Delimiter
                    DILUTION FACTOR
                    Delimiters
                    LEVEL
                    Delimiter
                    RECORD SEQUENCE NO.
                    CHECKSUM
FORMAT/CONTENTS
"22"
I
YYYY/blank (for PEST)15
I
MM/blank (for PEST)
I
DD/blank (for PEST)
I
HH/blank (for PEST)
I
MM/blank (for PEST)
I
Character/blank (for PEST)16
INI
Numeric/blank (for VOA)17
I
Numeric18
III
Numeric/blank (for VOA/SV)19
I
Numeric
Character
       15For volatiles and semivolatiles,  enter the date and time of analysis
of the most  recent  5  ug/L (VOAs)  or  the 20  ng/uL  (SVs)  standard  run prior  to
the sample reported in the  associated type  20  record.   Leave blank  for
pesticides.

       16Lab File ID of standard specified in 15 above (volatiles/semivolatiles
only).  This field  must match the Lab File  ID  on  Type  21  for the associated
calibration  (VSTD005/SSTD020).   Leave blank for pesticides.

       17Enter the Concentrated Extract Volume  for all SV and PEST.   The value
should be reported  in microliters.

       "Dilution factor of sample analyzed.

       "Concentration level of Pesticide  Individual Mix A and B  standards.
Concentration of low  point,  mid point and high point calibration standards as
a multiplier of  low point.   Low point = 1.0; Mid  point =  4.0;  High  point •
16.0.
OLC03.2
                                           H-18

-------
                                                                   Exhibit H -- Section  9
                                                                   Record Listing  (Con't)
9.6    Associated Injection and Counter Record (Type 23)

       Use:       Continuation of  type  20. Used to  identify associated blanks
                  and tunes, and the number of surrogates/DMCs and  spikes
                  outside  of the Quality  Control  (QC)  limits and the number of
                  tentatively identified  compounds  (TICs).
       Position:  Follows  the type 20,  21, and 22 to which, it applies.
        MAXIMUM
        LENGTH      CONTENTS
            2       RECORD TYPE
            1       Delimiter
            1       INSTRUMENT PERFORMANCE
                    CHECK (IPC/TUNE)  LABEL
            1       Delimiter
            4       IPC/TUNE INJECTION YEAR
            1       Delimiter
            2       IPC/TUNE INJECTION
                    MONTH
            1       Delimiter
            2       IPC/TUNE INJECTION DAY
            1       Delimiter
            2       IPC/TUNE INJECTION HOUR
            1       Delimiter
            2       IPC/TUNE INJECTION
                    MINUTE
            1       Delimiter
           14       DFTPP/BFB LAB FILE ID
            1       Delimiter
            2       VOLATILE STORAGE BLANK
                    LABEL
            1       Delimiter
            4       STORAGE BLANK INJECTION
                    YEAR
            1       Delimiter
            2       STORAGE BLANK INJECTION
                    MONTH
            1       Delimiter
            2       STORAGE BLANK INJECTION
                    DAY
            1       Delimiter
            2       STORAGE BLANK INJECTION
                    HOUR
            1       Delimiter
            2       STORAGE BLANK INJECTION
                    MINUTE
            1       Delimiter
           14       STORAGE BLANK LAB
                    FILE  ID  (VOA ONLY)
            4       Delimiters
FORMAT/CONTENTS
"23"
I
"P" (for BFB and DFTPP IPC) or blank
(for pesticides)
I
YYYY/blank (for PEST)
I
MM/blank (for PEST)

I
DD/blank (for PEST)
I
HH/blank (for PEST)  .
I
MM/blank (for PEST)

I
Character/blank  (for PEST)
I
"HB"  (fpr VOA) or blank  (for SV and
PEST)
I
YYYY/blank (for SV and PEST)

I
MM/blank (for SV and PEST)

I
DD/blank (for SV and PEST)

I
HH/blank (for SV and PEST)

I
MM/blank (for SV and PEST)

I
Character
                                           H-19
                                                                                   OLC03.2

-------
Exhibit H -- Section 9
Record Listing  (Con't)
        Associated Injection and Counter Record  (Type  23)  (Con't)
        MAXIMUM
        LENGTH      CONTENTS
            2       METHOD BLANK LABEL

            1       Delimiter
            4       METHOD BLANK INJECTION
                    YEAR
            1       Delimiter
            2       METHOD BLANK INJECTION
                    MONTH
            1       Delimiter
            2       METHOD BLANK INJECTION
                    DAY
            1       Delimiter
            2       METHOD BLANK INJECTION
                    HOUR
            1       Delimiter
            2       METHOD BLANK INJECTION
                    MINUTES
            1       Delimiter
           14       METHOD BLANK LAB
                    FILE (for VOA and
                    SV)/SAMPLE ID (for
                    PEST)
            1 •      Delimiter
            1       SURROGATE (for PEST)/
                    DMC  (for VOA and SV)
                    RECOVERY LABEL
            1       Delimiter
            2       SURROGATE (for PEST)/
                    DMC  (for VOA and SV)
                    RECOVERIES OUT
            1       Delimiter
            1       TIC LABEL

            1       Delimiter
            2       NO. OF TICS
            1       Delimiter
            1       SPIKE RECOVERY LABEL

            1       Delimiter
FORMAT/CONTENTS
"MBn/blank  (for standard, tune, and
method blanks)
I
YYYY/blank  (for standard, tune, and
method blanks)
I
MM/blank  (for standard, tune, and method
blanks)
I
DD/blank  (for standard, tune, and method
blanks)
I
HH/blank  (for standard, tune, and method
blanks)
I
MM/blank  (for standard,.tune, and method
blanks)
I
CHARACTER
 "P" for % recoveries/blank  (for STD/IPC)
Numeric20
"T"  (for VOA and SV TICs)/blank  (for
PEST)
I
Numeric
I
"S" for LCS (Pest)/MS/MSD  (all
fractions)/blank for anything else
       20This will be the number of surrogate (for PEST) or DMC (for VOA and
SV)  recoveries  outside QC limits for a specific column.  It should not be
cumulative  of the two columns for pesticides.
OLC03.2
                                           H-20

-------
                                                                   Exhibit H -- Section 9
                                                                   Record Listing (Con't)
       Associated Injection and Counter Record  (Type 23)  (Con't)

       MAXIMUM
       LENGTH      CONTENTS                     FORMAT/CONTENTS
           2       SPIKE RECOVERIES OUT         Numeric/blank21
           1       Delimiter                    |
           1       RPD LABEL                    "R" for RPD/blank22
           1       Delimiter                    |
           2       RPD OUT                      Numeric
           1       Delimiter                    |
           5       RECORD SEQUENCE NO.          Numeric
           4       CHECKSUM                     Character
       21Enter the number of  spike  recoveries  out.   Enter  "0"(zero)  if  none  of
the spike recoveries are outside of the QC limit.

       22"R"  for Matrix Spike/Matrix Spike Duplicate Recovery  Relative  Percent
Difference.  Leave blank for all other samples.

                                           H-21                                   OLC03.2

-------
Exhibit H -- Section 9
Record Listing  (Con't)

9.7    Sample Cleanup Record (Type 27)

       Use:        Continuation of  type  20. Used  to  identify sample/blank cleanup
                  procedures and QC  results.
       Position:  Follows  type 20, 21,  22, and 23 to which it applies.
        MAXIMUM
        LENGTH      CONTENTS
            2       RECORD TYPE.
            8       Delimiters
            1     .  FLORISIL CLEANUP TYPE
            1       Delimiter
            4       FLORISIL LOT  CHECK YEAR
            1       Delimiter
            2       FLORISIL LOT  CHECK MONTH
            1       Delimiter
            2       FLORISIL LOT  CHECK DAY
            1       Delimiter
            2       FLORISIL LOT  CHECK HOUR
            1       Delimiter
            2       FLORISIL LOT  CHECK MINUTE
            1       Delimiter
           14       FLORISIL DATA DESCRIPTOR
            1       Delimiter
            1       SULFUR CLEANUP

            1       Delimiter
            2       SULFUR BLANK  LABEL
            1       Delimiter
            4       SULFUR BLANK INJECTION YEAR
            1       Delimiter
            2       SULFUR BLANK INJECTION MONTH
            2       Delimiters
            2       SULFUR BLANK INJECTION DAY
            1       Delimiter
            2       SULFUR BLANK INJECTION HOUR
            1       Delimiter
            2       SULFUR BLANK INJECTION MINUTE
            1       Delimiter
           14       SULFUR BLANK LABORATORY/
                    SAMPLE ID
            1       Delimiter
            5       RECORD SEQUENCE NO.
            4       CHECKSUM
FORMAT/CONTENTS
"27"
mini
"F"/blank (for VOA and SV)
I
YYYY/blank (for VOA and SV)
I
MM/blank (for VOA and SV)
I
DD/blank (for VOA and SV)
I
HH/blank (for VOA and SV)
I
MM/blank (for VOA and SV)
I
Character23
I
Y/N (for PEST)/blank (for VOA
and SV)
I
"SB"/blank (if no separate
sulfur blank was prepared for
pesticides; also blank for VOA
and SV)
I
YYYY/blank (for VOA and SV)
I
MM/blank (for VOA and SV)
II
DD/blarik (for VOA and SV)
I
HH/blank (for VOA and SV)
I
MM/blank (for VOA and SV)
I
Character
Numeric
Character
       "Lab Sample ID of associate Florisil lot check.   This  is  a unique
 identifier  assigned to a lot  of  Florisil  cartridges.
OLC03.2
                                           H-22

-------
       Results Data Record  (Type 30)  (Con't)
                                                                   Exhibit  H --  Section 9
                                                                   Record Listing (Con't)
       MAXIMUM
       LENGTH
           1
CONTENTS
Delimiter
                                  FORMAT/CONTENTS
                   MEAN CONCENTRATION LABEL
           1       Delimiter
          13       MEAN CONCENTRATION

           1       Delimiter
           1       PERCENT DIFFERENCE LABEL

           1       Delimiter
           5       PERCENT DIFFERENCE
           1       Delimiter
           1       INTERNAL STANDARD AREA LABEL

           1       Delimiter
          13       INTERNAL STANDARD AREA

           1       Delimiter
           5       RECORD.SEQUENCE NO.
           4       CHECKSUM
                                   "M"  for Mean cone,  (for
                                   multicomponent  PEST only)/blank
                                   (for VOA  and SV)
                                   I
                                   Numeric  (for PEST)/blank (for VOA
                                   and  SV)32
                                   I
                                   "F"  or "P"  (PEST)/blank (for VOA
                                   and  SV field sample analysis)33
                                   I       '•    .
                                   Numeric
                                   I
                                   "I"  for  IS  Area (for VOA and
                                   SV)/blank (for  PEST)
                                   I
                                   Numeric  (for VOA and SV)/blank
                                   (for PEST)
                                   I
                                   Numeric
                                   Character
       32Mean Concentration  for Multicomponent  analytes  detected  in pesticide
analyses.

       33npn for percent Difference  between  concentrations  from two columns  in
pesticide analyses, or "F"  for Percent Difference between average  RRF  (initial
calibration) and RRF5/RRF20  (continuing calibration) in VOA/SV analyses.
Leave blank for volatile and semivolatile sample, blank, and tune  analysis.
                                           H-25
                                                                                  OLC03.2

-------
Exhibit H -- Section 9
Record Listing  (Con't)

9.9    Auxiliary Data Record  {Type 32)

       Use:         Used to report  retention time (in minutes)  for Internal
                   Standards and for  TICs (for Volatiles and Semivolatiles).  Used
                   to. report retention- time data and percent breakdown (for
                   pesticides).
       Position:   Follows type  30.  (Record will only be required as specified
                   above.)
        MAXIMUM
        LENGTH      CONTENTS
            2       RECORD TYPE
            3       Delimiters
            2       RETENTION TIME LABEL
            1       Delimiter
            5       RETENTION TIME
            1       Delimiter
            3       FIRST LIMIT LABEL
            1       Delimiter
            5       RT WINDOW LOWER LIMIT
            1       Delimiter
            3       SECOND LIMIT LABEL
            1       Delimiter
            5       RT WINDOW UPPER LIMIT
            2       Delimiters
            2       % BREAKDOWN LABEL

            1       Delimiter
            5       % BREAKDOWN
            1       Delimiter
            5       COMBINED % BREAKDOWN
            2       Delimiters
            1       PEAK
            1       Delimiter
            5       RECORD SEQUENCE NO.
            4       CHECKSUM
FORMAT/CONTENTS
"32"
III
"RT"
I
Numeric
I
"RTF"
I
Numeric
I
"RTT"
I
Numeric
II
"PB" for % breakdown/blank  (for
VOA and SV)
I
Numeric (DDT/ENDRIN)/blank  (for
VOA and SV)

I
Numeric/blank (for VOA and SV)34
II
1 THROUGH 5 (for pesticide
multicomponent compounds)/blank
(for VOA and SV) 3S
I
Numeric
Character
       34The combined %breakdown will be reported on both the record type 32s
 for DDT and Endrin.
       35For positively identified compounds,  a minimum of 3 peaks and a
maximum of  5 peaks  are allowed.   Types  30  and 32  will  be  repeated for each
peak that  is reported (a minimum of  three,  a maximum of  five  times).
for multicomponent  analytes  in pesticide  analyses.
                This is
OLC03.2
                                           H-26

-------
                                                                   Exhibit H -- Section 9
                                                                   Record Listing (Con't)
9.10   Name Record (Type 33)
       Use:        This record type is used for volatile and semivolatile
                  analyses only to carry an analyte name for TICs.   This record
                  is not used for pesticide analysis.
       Position:   Follows types 30 and 32 for TICs.

        MAXIMUM
        LENGTH      CONTENTS                           FORMAT/CONTENTS
            2        RECORD TYPE                        "33"
            1       .Delimiter                          |
          67        NAME OF COMPOUND                   Character
            1        Delimiter                          |
            5        RECORD SEQUENCE NO.                Numeric
            4        CHECKSUM                           Character
                                           H-27                                   OLC03.2

-------
Exhibit H -- Section 9
Record Listing  (Con't)

9.11   Instrumental Data Readout Record  (Type 36)
       Use:
       Position:

        MAXIMUM
        LENGTH
            2
            1
            1   •
            3
            3
            2
            5

            1
            3
            1
            5

            1
            5
            1
            3
            1
            5

            2
            3
            1
            5

            1
            5
            1
            3
            1
            5

            1
            5
            1
            3
            1
This record type is only used for volatile and semivolatile
analyses to describe decafluorotriphenylphosphine/4-
bromofluorobenzene (DFTPP/BFB) percent abundances.  This
record is not used for pesticide analysis.
Follows type 30 for DFTPP/BFB data.
 CONTENTS
 RECORD TYPE
 Delimiter
 MASS LABEL
 Delimiters
 FIRST MASS  (DFTPP/BFB)
 Delimiters
 FIRST PERCENT RELATIVE
 ABUNDANCE
 Delimiter
 SECOND MASS
 Delimiter
 SECOND PERCENT RELATIVE
 ABUNDANCE
 Delimiter
 PERCENT MASS OF 69
 Delimiter
 THIRD MASS
 Delimiter
 THIRD PERCENT RELATIVE
 ABUNDANCE
 Delimiters
 FOURTH MASS
 Delimiter
 FOURTH PERCENT RELATIVE
 ABUNDANCE
 Delimiter
 PERCENT MASS OF 69
 Delimiter
 FIFTH MASS
 Delimiter
 FIFTH PERCENT RELATIVE
 ABUNDANCE
 Delimiter
 PERCENT MASS OF 174
 Delimiter
 SIXTH MASS
 Delimiter
FORMAT/CONTENTS
"36"
III
Numeric  (DFTPP for SV or BFB for VOA)
II
Numeric

I  -
Numeric
I
Numeric

I
Numeric, DFTPP only/blank  (for VOA)
I
Numeric
I
Numeric

II                            '
Numeric
I
Numeric

I
Numeric, DFTPP only/blank  (for VOA)
I
Numeric
I
Numeric

I
Numeric, BFB only/blank (for SV)
I
Numeric
OLC03.2
                                           H-28

-------
                                                           Exhibit H -- Section 9
                                                           Record Listing (Con't)
Instrumental Data Readout Record (Type 36} (Con't)
MAXIMUM
LENGTH     CONTENTS
    5       SIXTH  PERCENT RELATIVE
           ABUNDANCE
    2       Delimiters
    3       SEVENTH MASS
    1       Delimiter
    5       SEVENTH PERCENT RELATIVE
           ABUNDANCE
    1       Delimiter
    5       PERCENT MASS OF 174
    1       Delimiter
    3       EIGHTH MASS
    1       Delimiter
    5       EIGHTH PERCENT RELATIVE
           ABUNDANCE
    1       Delimiter
    5        PERCENT MASS OF 174
    1       Delimiter
    3       NINTH MASS
    1       Delimiter
    5       NINTH PERCENT RELATIVE
            ABUNDANCE
    1        Delimiter
    5        PERCENT MASS OF 176
    1        Delimiter
    3        TENTH MASS
    '1       Delimiter
    5       TENTH PERCENT RELATIVE
            ABUNDANCE
    2       Delimiters
    3       ELEVENTH MASS
    1       Delimiter
    5       ELEVENTH  PERCENT
            RELATIVE ABUNDANCE
    2    •   Delimiters
    3       TWELFTH MASS
    1       Delimiter
    5       TWELFTH  PERCENT RELATIVE
            ABUNDANCE
    2       Delimiters
    3       THIRTEENTH  MASS
    2       Delimiters
    5       THIRTEENTH  PERCENT
            RELATIVE ABUNDANCE
    1       Delimiter
    5       PERCENT  MASS OF 442
    1       Delimiter
    5       RECORD SEQUENCE NO.
    4       CHECKSUM
FORMAT/CONTENTS
Numeric

II
Numeric
I
Numeric

I
Numeric, BFB only/blank  (for SV)
I
Numeric
I
Numeric
Numeric, BFB only/blank  (for SV)
 I
Numeric
 I
Numeric
 Numeric,  BFB  only/blank (for  SV)
 I
 Numeric/blank (for VGA)
 I
 Numeric/blank (for VOA)

 II
 Numeric/blank (for VOA)
 I
 Numeric/blank (for VOA)

 II
 Numeric/blank (for VOA)
 I
 Numeric/blank (for VOA)

 II
 Numeric/blank (for VOA)
 II
 Numeric/blank (for VOA)

 I
 Numeric, DFTPP only  (blank for VOA)
 I
 Numeric
 Character
                                    H-29
                                                                           OLC03.2

-------
Exhibit H -- Section 9
Record Listing  (Con't)

3.12   Comment Record (Type 90)

       Use:       To provide  for  operator-entered comments.
       Position:  May occur anywhere  in  the  file  after the  type  10  record.

        MAXIMUM
        LENGTH      CONTENTS                           FORMAT/CONTENTS
            2       RECORD TYPE                       "90"
            1       Delimiter                           |
           67       ANY  COMMENT                       Character
            1       Delimiter                           |
            5       RECORD SEQUENCE NO.                Numeric
            4       CHECKSUM                           Character
OLC03.2                                    H-30

-------
                                                                •  Exhibit H -- Section 10
                                                             Definitions of Various Codes
10.0   DEFINITIONS OF VARIOUS CODES USED IN AGENCY STANDARD RECORDS

10.1   Quality Control and Related Codes (QCC)  in Type 20 Records

10.1.1    NOTE: These codes appear in the quality control  (QC) code fields of
          type 20 records.  They are used to indicate the type of data that is
          being reported.
          OCC   Name

          LRB   LABORATORY  (REAGENT)
                BLANK

          LIB   LABORATORY  INSTRUMENT
                BLANK

          LSB   LABORATORY  SULFUR BLANK
           LHB   LABORATORY STORAGE BLANK

           FRB   FIELD BLANK
           FRM  FIELD REFERENCE SAMPLE
Definition

The "Method Blank" (Exhibit G).


The "Instrument Blank".


If different from "Method Blank"
(pesticides).

The storage blank (volatiles).

This is any sample that is submitted
from the field and is identified as
a blank.  This includes trip blanks,
rinsates, equipment blanks, etc.

This is any sample that is submitted
for a Case and is identified as a
Performance Evaluation  (PE) sample.
           LSD  LABORATORY SPIKE
                DUPLICATE BACKGROUND
                 (ORIGINAL) VALUES
           LF1
                LABORATORY SPIKE SAMPLE
                  FINAL   FIRST MEMBER
           LF2   LABORATORY SPIKE SAMPLE
                  FINAL  SECOND MEMBER
An environmental sample which is
analyzed according to the analytical
method, and subsequently used for
the matrix spike, and the matrix
spike duplicate  (Exhibit G).

The "Matrix spike"  (Exhibit G); must
precede LF2.

The "Matrix spike Duplicate"
 (Exhibit G).
           LCM   LABORATORY CONTROL SAMPLE
                                               The  Laboratory Control  Sample
                                                (Exhibit  G).
                               (LCS)
           LPC   LABORATORY PERFORMANCE
                 CHECK SOLUTION
A solution of  DFTPP  (SV)  or BFB
 (VOA)  or method  analytes  (PEST/PCB)
used to evaluate the  performance  of
an instrument  with respect  to a
defined set  of criteria (Tune or
Resolution Check Sample)  (Exhibit
G) .
           FLO   FLORISIL CHECK SOLUTION
A solution of pesticides used to
check recovery from each lot of
Florisil  cartridges.   These recovery
results will be provided in every
production run where associated
samples are analyzed.
                                           H-31
                                                                                   OLC03.2

-------
Exhibit H -- Section 10
Definitions of Various Codes  (Con't)
           OCC  Name

           CLM  INITIAL  CALIBRATION -
                MULTI-POINT
           CLS   INITIAL CALIBRATION
                 SINGLE POINT
           CLC   CONTINUING CALIBRATION
                 CHECK

           CLE   CONTINUING PERFORMANCE
                 CHECK
           CLD   DUAL PURPOSE CALIBRATION
  Definition

  The Initial Calibration for GC/MS
   (Exhibit G), or the Initial
  Individual Standard Mixes  (A, B) for
  pesticides  (Exhibit D PEST).
  Response factors  (GC/MS) or
  Calibration Factors  (pesticides)
  will be reported on the following
  type 30 records.

  The Initial Toxaphene/Aroclor Mixes
  used to determine all calibration
  factors  (Exhibit D PEST).

  The continuing calibration  (VSTD005/
  SSTD020) for GC/MS.

  The subsequent Individual Standard
  Mixes  (A,B), Performance Evaluation
  Mixture, and for subsequent
  injections of Toxaphene/Aroclor
  mixes  for pesticides  (Exhibit D
  PEST).

  A calibration solution as above used
  both as an initial calibration  (CLM)
  and a  continuing check  (CLC).
10.1.2     The following QCC values are used on type 20 records which act as a
           header, and indicate that additional (usually calculated) analyte
      *    specific data will be present on type 30 (and following type)
      •'-'    records.  Usually, these data will apply to an entire production run,
      •:-'    in which case they will appear immediately following the type 10
           record or type 11 record if present.  If the data apply to only a
           portion of the samples in the run, they shall be placed immediately
           preceding the samples to which they apply.   Much of the rest of the  .
           information in the type 20 record may be blank, indicating that these
           data do not apply to these results.
           MNC
                  MEAN VALUES FROM
                  CALIBRATIONS
The data following represent mean
values and percent RSDs from the
initial calibration (GC/MS) or the
mean calibration factors,  mean
retention times, and retention
time windows (pesticides).
OLC03.2
                                           H-32

-------
                                                                  Exhibit H -- Section 10
                                                     Definitions of Various Codes (Con't)
10.2   Codes For Sample Medium (Matrix,  Sources)
             Medium                                             Code
                All Media,  Specific Medium not Applicable.       0 (zero)
                      Use for Calibrations,  Tunes,  etc.
             Water                                              1

10.3   List of  Sample and Result Qualifiers

       Definition:   A sample qualifier consists  of three characters which act
       as an indicator of the fact and the reason that the subject analysis (a)
       did not  produce a numeric result,  or (b)  produced a numeric result for
       an entire sample but it is qualified in some respect relating to the
       type or  validity of the result.

10.3.1     Sample Qualifiers


           Qualifier     Full Name       Definition

           RIN            RE-ANALYZED     The indicated analysis results were
                                        generated from a re-injection of the same
                                        sample extract or aliquot  (RE SUFFIX).

           REX            RE-PREPARED     The indicated analysis results were
                                        generated from a re-extraction of the same
                                        sample  (RE SUFFIX).

           REJ            REJECTED        The results for the  entire sample analysis
                                        have been rejected for an unspecified
                                        reason by the laboratory.  For initial
                                        calibration data, these  data were not
                                        utilized in the calculation  of the mean.

           SRN            DILUTED        The indicated analysis results were
                                        generated from a dilution of the same
                                        sample  (DL SUFFIX).

10.3.2     Result Qualifiers in Type 30 Records

           A result qualifier consists of  three  characters  which  act  as an
           indicator of the fact and the reason  that the  subject  analysis (a)
           did  not  produce a numeric result, or  (b) produced  a numeric  result
           for  a single analyte but it is  qualified in some respect  relating  to
           the  type or validity of the result.   This qualifier is complementary
           to the  flags field on a type 30 record.  A  tentatively identified
           compound (TIC) must have either a TIE, TFB,  or PRE result  qualifier.
           BDL   BELOW DETECTABLE           Indicates compound was analyzed for but
                 LIMITS                     not detected  (Form 1 "U" Flag).

           NAR   NO ANALYSIS RESULT         There is no analysis result required
                                            for this subject parameter.

           AVG   AVERAGE VALUE              Average value -- used to report a range
                                            of values  (e.g., relative response
                                            factors).

           CBC   CANNOT BE CALCULATED       The analysis result cannot be
                                            calculated because an operand value is
                                            qualified  (e.g., identifies analytes
                                            whose internal standard is not found)
                                            (Form 1  "X" Flag).
                                           H-33                                    OLC03.2

-------
Exhibit H -- Section 10
Definitions of Various Codes  (Con't)
           LTL   LESS THAN LOWER
                CALIBRATION LIMIT
           GTL   GREATER THAN UPPER
                CALIBRATION LIMIT


           LLS   LESS THAN LOWER
                STANDARD
           TIE  TENTATIVELY  IDENTIFIED
                ESTIMATED VALUE
           REJ  REJECTED
           STD  INTERNAL  STANDARD
           STB   INTERNAL  STANDARD
                 BELOW  DETECTION LIMITS

           FBK   FOUND  IN  BLANK
           TFB  TENTATIVELY IDENTIFIED
                AND  FOUND IN BLANK

           NRP  NON-REPRODUCIBLE
           PRE   PRESUMPTIVE PRESENCE
           ALC  ALDOL CONDENSATION
Analysis result is  from a diluted
sample  (DL suffix)  and may be less
accurate than the result from an
undiluted sample  (Form 1 "D" Flag).

Actual value is known to be greater
than the upper calibration range  (Form
1  "E" Flag).

The analysis result is less than the
sample quantitation limit  (Form 1  "J"
Flag).

The indicated analyte is a tentatively
identified analyte; its concentration
has been estimated  (Form 1LCF or 1LCG
"J" Flag).

Results for the analyte are rejected by
the laboratory.

The indicated compound is an internal
standard.

A  combination of  "STD" and "BDL".
 The  indicated compound was  found  in  the
 associated method blank (LRB)  as  well
 as the  sample (Form 1  "B" Flag).
A Combination of  "TIE"  and
'1LCF  or  1LCG "B"  Flag).
'FBK"   (Form
 Results  of  two or more injections  are
 not  comparable (Form 1LCE  "P"  flag),
 e.g., Aroclor target analyte with
 greater  than 25%  difference between
 mean concentrations of the two column
 analyses.

 Presumptive evidence of presence of
 material for TIC  (Form 1LCF or 1LCG
 "N"  Flag).

 Labels a suspected Aldol Condensation
 product  for TICs  (Form 1LCG)
OLC03.2
                                           H-34

-------
PURCHASE CARD LOG FOR BILLING PERIOD: May 1 - 31, 2001
                                                PAGE 1 OF 2
PRINTED NAME OF
PURCHASE CARDHOLDER: Mary S. Price
ACCOUNT NUMBER: 4486 8600 0014 0972
ITEMS
PURCHASED
Plastic inserts for ID'S (500)
Toner for Fax (E126/A224)
Floor Tread/Sealer/Floor Prep
Safety Signs/labels
Battery for 2-way radio (John J.)
Overhead bulbs (FXL)/labels
Repair signaling module on fire pump
Cartridges for Label Maker used for
name plates
Safety Glasses (Luisa S.)
V
Battery for 2-way Radio (Skip)
Whiteboard (A235)
VENDOR/
MERCHANT
Identicard
Dove Data
Lab Safety
Lab Safety
Townex Communications
Mali
Honeywell
Mali
Unicor
Townex Communications
Herman Miller
APPROVALS
(IF APPROPRIATE)
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
DATE
ORDERED
05/02/01
05/02/01
05/03/01
5//7/01
05/07/01
05/10/01
05/14/01
05/14/01
05/15/01
05/22/01
05/22/01
TOTAL
COST
$148.43
$116.70
$606.39
$438.54
$69.00
$109.95
$305.00
$78.32
$53.00
$116.00
$92.54
OBJECT
CLASS
26.15
26.15
25.82
25.82
31.48
26.15
25.82
26.15
26.29
31.48
31.43
DATE RECONCILED
IN EAGLS
05/09/01
05/11/01
05/11/01
05/11/01
05/17/01
05/17/01
06/04/01
05/21/01

06/04/01

SIGNATURE OF SIGNATURE OF
PURCHASE CARDHOLDER: APPROVING OFFICIAL:
DEFAULT
DCN*
0103JM0013
0103JM0010
0103JM0010
0103JM0010
0103JM0013
0103JM0010
0103JM0013
0103JM0013
0103JM0012
0103JM0013
0103JM0010



-------
PURCHASE CARD LOG FOR BILLING PERIOD:  May 1-31, 2001                                                         PAGE 2 OF 2



                                                                                 ACCOUNT NUMBER: 4486860000140972
FINANCIAL DATA FOR PURCHASES NOT CHARGEABLE TO CARDHOLDER'S "DEFAULT DCN"

I    DON    I   BUDGET  I     FUND     I     ORG     I     PRO    I       SITE7PROJ      I   COST/ORG   I  OBJ CLASS  I     AMOUNT
	I FISCAL YEAR I              |	|	   I                    I	I	\	

-------
PURCHASE CARD LOG FOR BILLING PERIOD: May 1-31, 2001
                                  PAGE 1 OF 2
PRINTED NAME OF
PURCHASE CARDHOLDER: Mary S. Price
ACCOUNT NUMBER: 4486 8600 0014 0972
ITEMS
PURCHASED
Sign for B202
Toner for Xerox copier (A1 01 )
E-Z Seal for Postage Meter








VENDOR/
MERCHANT
Environmental Siq
XEROX
Pitney Bowes








APPROVALS
(IF APPROPRIATE)
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE OF
PURCHASE CARDHOLDER:
TELEPHONE NUMBER: 410-305-2646
DATE:

DATE
ORDERED
05/25/01
05/25/01
05/31/01








TOTAL
COST
$60.00
$103.00
$93.09








OBJECT
CLASS
25.82
26.15
26.15





••


E RECONC
IN EAGLS











SIGNATURE OF
APPROVING OFFICIAL:
TELEPHONE NUMBER: 215-814-5 DATE:


DEFAULT
DCN*
0103JM0012
0103JM0013
0103JM0013
0103JMOO
0103JMOO
0103JMOO
0103JMOO
0103JMOO
0103JMOO
0103JMOO
0103JMOO
0003JM0009
0003JM0009
0003JM0009
0003JM0009
*PAGE 2 MUST BE COMPLETED IF ITEM PURCHASED WAS NOT CHARGED TO DEFAULT DCN/ACCOUNT NUMBER(S).

-------
                                                                                                       PAGE 1 OF 2
PRINTED NAME OF
PURCHASE CARDHOLDER: Mary S. Price
ACCOUNT NUMBER: 4486 8600 0000 9714
ITEMS
PURCHASED











VENDOR/
MERCHANT











APPROVALS
(IF APPROPRIATE)
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
SIGNATURE/TITLE
DATE
ORDERED











TOTAL
COST











OBJECT
CLASS











TE RECONCIL
IN EAGLS











SIGNATURE OF SIGNATURE OF
PURCHASE CARDHOLDER: APPROVING OFFICIAL:
DEFAULT
DCN*
0003JM0009
0003JM0009
0003JM0009
0003JM0009
0003JM0009
0003JM0009
0003JM0009
0003JM0009
0003JM0009





TELEPHONE NUMBER: 410-305-26 DATE:
TELEPHONE NUMBER: 215-814-   DATE:
•PAGE 2 MUST BE COMPLETED IF ITEM PURCHASED WAS NOT CHARGED TO DEFAULT DCN/ACCOUNT NUMBER(S).

-------