USEPA CONTRACT LABORATORY PROGRAM


                                      STATEMENT OF WORK

                                             FOR

                                     INORGANIC ANALYSIS


                              Multi-Media,  Multi-Concentration

                                           ILM05.2
                                        December  2001
Method 00060

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2

-------
                                           54OR01504
                 THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2

-------
                               STATEMENT OF WORK

                               TABLE OF CONTENTS



EXHIBIT A:  SUMMARY OF REQUIREMENTS

EXHIBIT B:  REPORTING AND DELIVERABLES REQUIREMENTS

EXHIBIT C:  INORGANIC TARGET ANALYTE LIST WITH CONTRACT REQUIRED QUANTITATION
            LIMITS

EXHIBIT D:  ANALYTICAL METHODS

EXHIBIT E:  CONTRACT LABORATORY PROGRAM QUALITY ASSURANCE MONITORING PLAN

EXHIBIT F:  CHAIN-OF-CUSTODY, DOCUMENT CONTROL AND WRITTEN STANDARD OPERATING
            PROCEDURES

EXHIBIT G:  GLOSSARY OF TERMS

EXHIBIT H:  DATA DICTIONARY AND FORMAT FOR DATA DELIVERABLES IN COMPUTER-
            READABLE FORMAT


APPENDIX A: FORMAT OF RECORDS FOR SPECIFIC USES

APPENDIX B: MODIFIED ANALYSIS
                                                                         ILM05.2

-------
       EXHIBIT A




SUMMARY OF REQUIREMENTS
          A-l                               ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               A-2

-------
                      Exhibit A - Summary of Requirements

                               Table of Contents

Section                                                                    Page


1.0   PURPOSE	5
 t

2.0   DESCRIPTION OF SERVICE   	   5

3.0   DATA USES	5

4.0   SUMMARY OF REQUIREMENTS 	   5
      4.1   Introduction to the Inorganic Statement of Work	5
      4.2   Overview of Major Task Areas   	6
                                      A-3                               ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               A-4

-------
                                                      Exhibit A — Sections 1-4
                                                                         Purpose
1.0   PURPOSE
      The purpose of the multi-media, multi-concentration inorganic analytical
      service is to provide analytical data for use by the U.S. Environmental
      Protection Agency (USEPA) in support of the investigation and clean-up
      activities under the Comprehensive Environmental Response, Compensation,
      and Liability Act of 1980 (CERCLA) and the Superfund Amendments and
      Reauthorization Act of 1986 (SARA).  Other USEPA Program Offices that
      have similar analytical data needs also use this service.

2.0   DESCRIPTION OF SERVICE

      The inorganic analytical service provides a contractual framework for
      laboratories.  This framework applies USEPA Contract Laboratory Program
      (CLP) analytical methods for the isolation, detection, and quantitative
      measurement of 23 metals (including mercury) and cyanide in water/
      aqueous and/or soil/sediment samples.  The analytical service contract
      provides specific contractual requirements by which USEPA will evaluate
      the data.

3.0   DATA USES

      This analytical service contract provides data which USEPA uses for a
      variety of purposes, such as: determining the nature and extent of
      contamination at a hazardous waste site, assessing priorities for
      response based on risks to human health and the environment, determining
      appropriate cleanup actions, and determining when remedial actions are
      complete.  The data may be used in all stages in the investigation of
      hazardous waste sites, including: site inspections, Hazard Ranking
      System  (HRS) scoring, remedial investigation/feasibility studies,
      remedial design, treatability studies, and removal actions.

      The data may also be used in litigation against Potentially Responsible
      Parties in the enforcement of Superfund legislation.  As a result, the
      Contractor must be aware of the importance of maintaining the integrity
      of the data generated under this contract, since it is used to make
      major decisions regarding public health and environmental welfare.  The
      Contractor may be required to appear and testify to the accuracy and/or
      validity of the data generated.

4.0   SUMMARY OF REQUIREMENTS

4.1   Introduction to the Inorganic Statement of Work

      The Statement of Work  (SOW) is comprised of eight exhibits and two
      appendices.  Exhibit A provides an overview of the SOW and its general
      requirements.  Exhibit B contains a description of the reporting and
      deliverables requirements, in addition to the data reporting forms and
      instructions.  Exhibit C specifies the Inorganic Target Analyte List
      (TAL) for this SOW with the Contract Required Quantitation Limits
      (CRQLs) for the sample matrices.  Exhibit D details the required
      analytical procedures to be used with this SOW and resulting contracts.
      Exhibit E provides descriptions of required Quality Assurance/Quality
      Control  (QA/QC), Standard Operating Procedures  (SOPs), QA/QC
      performance, and the reporting of data.  Exhibit F contains chain-of-
      custody and sample documentation requirements.  To ensure proper
      understanding of the terms utilized in this SOW, a glossary can be found
      in Exhibit G.  When a term is used in the text without explanation, the
      glossary meaning shall be applicable.  Specifications for reporting data
      in computer-readable format appear in Exhibit H.  Appendix A provides
      examples of the data format requirements specified in Exhibit H.
      Appendix B contains a description of the requirements for performing


                                      A-5                                ILM05.2

-------
Exhibit A — Section 4
Summary of Requirements  (Con't)

      modified analyses, as well as the analytical procedure for Graphite
      Furnace Atomic Absorption  (GFAA).

4.2   Overview of Major Task Areas

      For each sample, the Contractor  shall perform the tasks described in
      each section.  Specific  requirements for each task are detailed in the
      exhibits referenced in the following sections.

4.2.1    Task I: Sample Receiving, Storage, and Disposal

4.2.1.1     Chain-of-Custody

            The Contractor  shall receive and maintain  samples under proper
            chain-of-custody.  All associated document control  and inventory
            procedures  shall be  developed and followed.  Documentation
            described herein shall be  required to  show that all procedures are
            strictly followed.  This documentation shall be reported as the
            Complete Sample Delivery Group  (SDG) File  (CSF)  (see Exhibit B).
            The Contractor  shall establish and use appropriate  procedures to
            safeguard confidential information received  from USEPA.

4.2.1.2     Sample  Scheduling/Shipments

            Sample  shipments to  the Contractor's facility  will  be scheduled
            and coordinated by the Contract Laboratory Program  (CLP) Sample
            Management  Office  (SMO).   USEPA may request  analyses that include
            all or  a  subset of the Inorganic Target Analytes listed in Exhibit
            C.  The Contractor shall  communicate with  SMO  personnel by
            telephone as necessary throughout the  process  of sample
            scheduling,  shipment, analysis, and data reporting, to ensure that
            samples are properly processed.

4.2.1.2.1      Samples  will be shipped routinely to the  Contractor through an
                overnight delivery  service.  However, as  necessary, the
                Contractor shall  be  responsible  for any handling or processing
                of the  receipt  of sample shipments. This includes  the pick-up
                of samples at  the nearest servicing airport,  bus station,  or
                other  carrier  within  the Contractor's geographical  area.   The
                Contractor shall  be  available  to  receive  and process  sample
                shipments at any  time  the delivery  service  is operating,
                including Saturdays,  to ensure  that short sample analysis  time
                requirements can  be  met.

4.2.1.2.2       If there are problems  with  the samples  (e.g., mixed media,
                containers broken or leaking)  or  sample documentation  and
                paperwork (e.g.,  Traffic Reports/Chain  of Custody Records  not
                with shipment,  sample  and Traffic  Report/Chain  of Custody
                Record do not correspond),  the Contractor shall  immediately
                contact SMO for resolution.   The  Contractor shall immediately
                notify SMO and the USEPA Regional  CLP  Project Officer  (CLP PO)
                regarding any problems and laboratory  conditions that  affect
                the timeliness of analyses and data reporting.   In particular,
                the Contractor shall immediately notify SMO personnel  and  the
                USEPA Regional CLP PO in advance  regarding sample data that
                will be delivered late and shall  specify the estimated delivery
                date.

4.2.1.2.3       To monitor the  temperature  of  the  sample  shipping cooler more
                effectively, each USEPA Regional  Office may include a sample
                shipping cooler temperature blank with each cooler shipped.
                The temperature blank will be clearly  labeled:  USEPA COOLER


 ILM05.2                               A-6

-------
                                                         Exhibit A  — Section  4
                                                Summary of Requirements  (Con't)

               TEMPERATURE  INDICATOR.  The Contractor  shall  record  the
               presence or  absence of the cooler  temperature indicator bottle
               on Form DC-1, Item 8 - Cooler Temperature Indicator  Bottle  (see
               Exhibit B).

4.2.1.2.3.1       When the USEPA Regional Office  supplies a  cooler  temperature
                  indicator bottle in the sample  shipping cooler, the
                  Contractor shall use the USEPA  supplied cooler temperature
                  indicator bottle to determine the  cooler temperature.  The
                  temperature  of the cooler shall be measured at the  time  of
                  sample  receipt by the  Contractor.

4.2.1.2.3.2       The temperature of the sample shipping cooler shall be
                  measured  and recorded  immediately  upon opening the  cooler.

4.2.1.2.3.3       To determine  the temperature of the  cooler:  the Contractor
                  shall locate the cooler temperature  indicator bottle in  the
                  sample  shipping cooler, remove  the cap, and insert  a
                  calibrated thermometer into  the cooler temperature  indicator
                  bottle.   Prior to recording  the temperature, the  Contractor
                  shall allow  a minimum  of 3 minutes,  but not greater than 5
                  minutes,  for the thermometer to equilibrate with  the liquid
                  in  the  bottle.  At a minimum, the  calibrated thermometer
                   (ą1°C)  shall have a measurable  range of 0-50°C.   Other
                  devices which can measure temperature may  be used if they
                  can be  calibrated to ą1°C and have a range of 0-50°C.  If a
                  temperature  indicator  bottle is not  present in the  cooler,
                  an  alternative means of determining  cooler temperature shall
                  be  used.  Under no circumstances shall a thermometer or  any
                  other device be inserted into a sample bottle for the
                  purpose of determining cooler temperature.  The Contractor
                  shall contact SMO and  inform them  that a temperature
                  indicator bottle was not present in  the cooler.   The
      -            Contractor shall document the alternative  technique used to
                  determine cooler temperature in the  SDG Narrative.

4.2.1.2.3.4       If the  temperature of  the sample shipping  cooler's
                  temperature  indicator  exceeds 10°C,  the Contractor  shall
                  contact SMO  and inform them  of  the temperature deviation.
                  SMO will  contact the Region  from which the samples  were
                  shipped for  instruction on how  to  proceed.  The Region will
                  either  require that no sample analysis(es)  be performed  or
                  that the  Contractor proceed  with the analysis(es).  SMO  will
                  in  turn notify the Contractor of the Region's decision.   The
                  Contractor shall document the Region's decision and the  EPA
                  sample  numbers of all  samples for  which temperatures
                  exceeded  10°C in the SDG Narrative.

4.2.1.2.3.5       The Contractor shall record  the temperature  of the cooler on
                  Form DC-1, under Item  9 - Cooler Temperature, and in the SDG
                  Narrative (see Exhibit B).

4.2.1.2.4      The Contractor  is required to retain  unused sample volume,  used
               sample containers, and empty sample bottle containers  for a
               period of  60 days after data submission.  From time  of receipt
               until  analysis,  the Contractor  shall  maintain all water/ao^jeous
                (preserved and  unpreserved) and/or soil/sediment samples  at 4°C
                (ą2°C)  (see  Exhibit B).

4.2.1.2.5      The Contractor  shall be required to routinely return sample
               shipping containers  (e.g., coolers) to  the appropriate sampling
                                      A-7                                ILM05.2

-------
Exhibit A — Section 4
Summary of Requirements  (Con't)

               office  within  14  calendar  days  following  shipment  receipt  (see
               contract,  Section G  titled,  "Government Furnished  Samples").

4.2.1.2.6      Sample  analyses will be scheduled by groups of samples, each
               defined as a Case and  identified by a unique  EPA Case  number
               assigned by SMO.   A  Case signifies a group of samples  collected
               at  one  site or geographical  area over a finite time  period, and
               will  include one  or  more field  samples with associated blanks.
               Samples may be shipped to  the Contractor  in a single shipment
               or  multiple shipments  over a period of time,  depending on  the
               size  of the Case.

4.2.1.2.6.1        A  Case consists of  one  or more SDGs.   An SDG  is defined by
                   the  following, whichever  is  most frequent:

                         Each Case of field  samples received, or

                         Each 20  field samples  [excluding Performance
                         Evaluation (PE)  samples]  within  a Case, or

                   •      Each 7 calendar day period  (3  calendar  day period for
                         7 day turnaround) during  which field samples in a Case
                         are received (said  period beginning with the receipt
                         of the first sample in the  SDG).

                   •      In addition,  all  samples  and/or  sample  fractions
                         assigned to an SDG  must have been scheduled under the
                         same contractual turnaround  time.  Preliminary Results
                         have no  impact on  defining the  SDG.

4.2.1.2.6.2        Samples may be assigned to SDGs by matrix  (i.e.,  all soils
                   in one SDG, all waters  in another),  at the discretion of the
                   laboratory.  However,  PE  samples  received within a Case
                   shall be assigned to an SDG  containing field  samples for
                   that Case.   Such assignment  shall  be made at  the time the
                   samples are received,  and shall not be made retroactively.

4.2.1.2.6.3        Each sample received by the  Contractor will be  labeled with
                   an EPA sample number,  and accompanied  by a Traffic
                   Report/Chain of Custody Record bearing the sample number and
                   descriptive information regarding  the  sample.  EPA sample
                   numbers are six digits  in length.   If  the Contractor
                   receives a sample number  of  any other  length, the Contractor
                   shall contact  SMO immediately.   The Contractor shall
                   complete and  sign the Traffic Report/Chain of Custody
                   Record, recording the date of sample receipt  and sample
                   condition on  receipt for each sample container.  The
                   Contractor shall also follow the instructions given on the
                   Traffic Report/Chain of Custody Record in choosing the
                   Quality Control  (PC) samples when such information is
                   provided.  If  no QC sample is designated on the Traffic
                   Report/Chain  of Custody Record, the Contractor shall select
                   a sample and  notify SMO  for Regional acceptance.  SMO shall
                   contact the Region for confirmation immediately after
                   notification.

 4.2.1.2.6.4        The Contractor shall submit  signed copies of  Traffic
                   Reports/Chain  of Custody Records for all samples in a SDG to
                   SMO within three working days following receipt of the last
                   sample  in the  SDG.  Faxed copies of Traffic Reports/Chain of
                   Custody Records do not meet this requirement.  Traffic
                   Reports/Chain  of Custody Records shall be submitted in SDG


 ILM05.2                               A-8

-------
                                                         Exhibit A — Section 4
                                                Summary of Requirements (Con't)

                  sets  (i.e., all Traffic Reports/Chain of Custody Records  for
                  a SDG shall be clipped together) with an SDG Cover Sheet
                  containing information regarding the SDG, as specified  in
                  Exhibit B.

4.2.1.2.6.5       EPA Case numbers, SDG numbers, and EPA sample numbers shall
                  be used by the Contractor  in  identifying samples received
                  under this contract both verbally and in reports/
                  correspondence.

4.2.1.3     Modified Analysis

            The Contractor may be requested by  USEPA to perform modified
            analyses.  These modifications will be within the scope of this
            SOW and may include, but are not limited to, analysis of
            additional analytes and/or lower quantitation limits.  These
            requests will be made by the USEPA  Regional CLP PO, USEPA OERR
            Analytical Operations/Data Quality  Center  (AOC) Inorganic Program
            Manager  (PM), and Contracting Officer  (CO) in writing, prior  to
            sample scheduling.  If the Contractor voluntarily elects to
            perform these modified analyses, these analyses will be performed
            with no increase in per sample price.  All  contract requirements
            specified in the SOW/Specifications will remain in effect unless
            the USEPA CO provides written approval for the modification (s) and
            a waiver for associated defects.  The USEPA CO approval must  be
            obtained prior to sample scheduling.

4.2.2    Task II: Sample Preparation and Analysis

4.2.2.1     Overview

            The Contractor is advised that the  samples received under this
            contract are usually from known  or  suspected .hazardous waste  sites
            and may contain high  (greater than  15%) levels of organic and
            inorganic materials of a potentially hazardous nature and of
            unknown structure and concentration, and should be handled
           • throughout the analysis with appropriate caution.  It is the
            Contractor's responsibility to take all necessary measures to
            ensure laboratory safety.

4.2.2.2     The Contractor shall prepare and analyze samples as described in
            Exhibit D.  Sample preparation methods shall remain consistent for
            all samples analyzed within a Case.  Prior to sample analysis, the
            Contractor shall review the Traffic Report/Chain of Custody Record
            for any special sample analysis  instructions.  Anomalies that
            occur during sample analysis shall  be reported to SMO and the
            USEPA Regional CLP PO immediately.

            The Contractor shall collectively review all analytical results
            associated with a sample.  This  includes undiluted, diluted,
            serial dilution, and interference results.  The Contractor shall
            report any significant anomalies between these results in the SDG
            Narrative indicating possible matrix interferences.

4.2.2.3     Quality Assurance/Quality Control Procedures

4.2.2.3.1      The Contractor shall strictly adhere to all specific QA/QC
               procedures prescribed in Exhibits D and E.  Records documenting
               the use of the protocol shall be maintained in accordance  with
               the document control procedures  prescribed in Exhibit F, and
               shall be reported in accordance  with Exhibits B and H.
                                      A-9                                ILM05.2

-------
Exhibit A — Section 4
Summary of Requirements  (Con't)

4.2.2.3.2      The Contractor  shall maintain a Quality Assurance Management
               Plan  (QAP)  with the objective of  providing sound analytical
               chemical  measurements.   This program shall incorporate  the QC
               procedures,  any necessary corrective action,  and all
               documentation required during data collection as well as  the
               quality assessment measures performed by  management to  ensure
               acceptable  data production.

4.2.2.3.3      Additional  QC shall be  conducted  in the form  of  the analysis  of
               laboratory  PE samples  submitted to the laboratory by USEPA.
               Unacceptable results of all such  QC or laboratory- PE samples
               may be  used as the basis for an equitable adjustment to reflect
               the reduced value of the data to  USEPA o.r_rejection of  the data
               for specific analyte(s)  within  an SDG or  the  entire SDG.  Also,
               unacceptable results may be used  as the basis for contract
               action.  "Compliant performance"  is defined as that which
               yields  correct analyte identification and concentration values
               as determined by USEPA,  as well as meeting the contract
               requirements for analysis (Exhibit D); QA/QC  (Exhibit E); data
               reporting and other deliverables  (Exhibits B  and H); and sample
               custody,  sample documentation,  and SOP documentation  (Exhibit
               F) .

4.2.3     Task  III: Sample  Reporting

4.2.3.1      USEPA has  provided to the Contractor formats for the reporting of
             data (Exhibits B and H).   The Contractor shall be responsible  for
             completing and submitting analysis data sheets,  computer-readable
             data on  diskette  (or via  an alternate means  of electronic
             transmission approved in  advance by  USEPA)  in a  format specified
             in this  SOW and within the time  specified in Exhibit B, Section
             1.1.

4.2.3.2      Use  of formats other than those designated by USEPA (see Exhibits
             B and H)  will be deemed  as noncompliant.  Such data are
             unacceptable.   Resubmission in the specified format at no
             additional cost to the Government  shall be required.

4.2.3.3      Computer generated forms  may be  submitted in the hardcopy  Sample
             Data Package (s) provided that the  forms are  in  exact  USEPA format.
             This means that the order of data  elements is the same  as  on each
             USEPA required form, including form numbers  and titles, page
             numbers, and header information.

4.2.3.4      The  data reported by the  Contractor  on the hardcopy data  forms and
             the  associated computer-readable data submitted by the  Contractor
             on diskette (or via an alternate means of electronic  transmission,
             if approved in advance by USEPA)  shall contain identical
             information.  If  discrepancies are found during Government
             inspection, the Contractor shall be required to resubmit  either
             the hardcopy forms or the computer-readable data, or both sets  of
             data, at no additional cost to USEPA.
 ILM05.2                               A-10

-------
               EXHIBIT B




REPORTING AND DELIVERABLES REQUIREMENTS
                   B-l                                ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               B-2

-------
              Exhibit B - REPORTING AND DELIVERABLES REQUIREMENTS

                               Table of Contents

Section                                                                    Page


1.0   CONTRACT REPORTS/DELIVERABLES DISTRIBUTION    	  5
      1.1   Report Deliverable Schedule  	   5
      1.2   Distribution   	   8

2.0   REPORTING REQUIREMENTS AND ORDER OF DATA DELIVERABLES 	   9
      2.1   Introduction   	   9
      2.2   Resubmission of Data   	9
      2.3   Quality Assurance  (QA) Management Plan  and
            Standard Operating Procedures  (SOPs)    	  10
      2.4   Sample Traffic Reports/Chain of Custody Records 	 10
      2.5   Sample Data Package	11
      2.6   Complete SDG File  (CSF )	16
      2.7   Data in Computer-Readable Format   	17
      2.8   Results of the Intercomparison and Performance
            Evaluation  (PE) Sample Analyse s 	  18
      2.9   Preliminary Results  	 18
      2.10  Quarterly Verification of Linear Ranges and Interelement
            Correction Factors and Annual Verification of MDLs    	  18
      2.11  Electronic Instrument Data   	19
      2.12  Corrective Action Procedures   	  19

3.0   FORM INSTRUCTIONS	20
      3.1   Introduction   	20
      3.2   General Information  	 20
      3.3   Header Information   	   20
      3.4   Inorganic Forms	22
      3.5   Sample Log-In Sheet  [Form DC-1]  	 46
      3.6   Full Inorganics Complete SDG File  (CSF) Inventory
            Sheet  [Form  DC-2]	48

4.0   DATA REPORTING FORMS   	49
                                       B-3                                ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               B-4

-------
                                                         Exhibit B — Section 1
                                     Contract Reports/Deliverables Distribution

1.0   CONTRACT REPORTS/DELIVERABLES DISTRIBUTION

1.1   Report Deliverable Schedule

      The following table reiterates the contract reporting and deliverables
      requirements and specifies the distribution that is required for each
      deliverable.  The turnaround times for Items B through E are 7, 14, or
      21 days.

      NOTE: Specific recipient names and addresses are subject to change
      during the term of the contract.  The USEPA Office of Emergency and
      Remedial Response  (OERR) Analytical Operations/Data Quality Center  (AOC)
      Inorganic Program Manager  (AOC PM) will notify the Contractor in writing
      of such changes when they occur.

                                    TABLE 1
Item
A.
B.2
C.2
D.2
E.2'3
F.Ť
G.5'6
Sample Traffic
Reports /Chain of
Custody Records
Sample Data Package
Data in Computer-
Readable Format
Results of
Intercomparison
Study /PE Sample
Analysis Study
Complete SDG File
(CSF)B
Preliminary Results
Quarterly
Verification of
ICP-AES/ICP-MS
Linear Ranges and
ICP-AES
Interelement
Correction Factors
Annual Verification
of Method Detection
Limits (MDLs)
No. of
Copies*
1
1
1
1
1
1
1
1
Delivery Schedule
3 working days after
receipt of last sample
in Sample Delivery
Group (SDG) -1
XXC days after
Validated Time of
Sample Receipt (VTSR)1
of last sample in SDG.
XXC days after VTSR of
last sample in SDG.
XXC days after VTSR of
last sample in SDG.
XXC days after VTSR of
last sample in SDG.
Within 72 hours after
receipt of each sample
at laboratory, if
requested.
Quarterly: 15th day of
January, April, July,
and October.
Annually: 15th day of
January.
Distribution
o
2
co
X
X
X
X

X
X
X
Region


X

X
X

o
O
0,
a,
u






X
X
to
H
a



X


X
X
                                      B-5
ILM05.2

-------
Exhibit B — Section 1
Contract Reports/Deliverables Distribution  (Con't)

                                TABLE  1  (Con't)


Item




H.6'7













I.6'7













J.







Standard Operating
Procedures (SOPs)












Quality Assurance
Management Plan
(QAP)











Electronic
Instrument Data







No. of
Copies*





1













1













Lot








Delivery Schedule





Revise within 30 days
after contract award
and receipt of USEPA
comments .

Submit within 7 days
of receipt of written
request to recipients
as directed. (See
Exhibit E, Section 6)
Submit within 14 days
of amended SOP(s) as
directed in Exhibit E,
Section 6.4.
Revise within 30 days
after contract award
and receipt of USEPA
comments .

Submit within 7 days
of receipt of written
request to recipients
as directed. (See
Exhibit E, Section 5)
Submit within 14 days
of amended QAP as
directed in Exhibit E,
Section 5.3.
Retain for 3 years
after data submission.
Submit within 7 days
after receipt of
written request by the
USEPA Regional CLP PO.
(See Exhibit E,
Section 13)
Distribution
Q
c o
O CU
H CO
O 0i Cu E-i
S Q) ťJ <
c/j os o a
As Directed

Amended SOPs
distributed to
CLP PO and QATS









As Directed

Amended QAP
distributed to
CLP PO and QATS









As Directed







 ILM05.2
                                       B-6

-------
                                                         Exhibit B — Section 1
                             Contract Reports/Deliverables  Distribution  (Con't)
Footnotes:
      AThe number of copies specified is the number of copies required to be
delivered to each recipient.

      Contractor-concurrent delivery to USEPA's designated recipient [e.g.,
Quality Assurance Technical Support  (QATS)] may be required upon request by
the USEPA OERR Analytical Operations/Data Quality Center  (AOC) Inorganic
Program Manager  (AOC PM).  Retain for 365 days after data submission, and
submit as directed within 7 days after receipt of written request by the USEPA
AOC PM.

      cThe number of days associated with these elements will be provided in
the associated laboratory contract document and will also be provided at the
time of sample scheduling by the Sample Management Office  (SMO) Contractor.

      "The CLP PO is the USEPA Regional Contract Laboratory Program (CLP)
Project Officer  (CLP PO) designated on the contract.

      Validated Time of Sample Receipt (VTSR)  is the date of sample receipt
at the Contractor's facility, as recorded on the shipper's delivery receipt
and sample Traffic Report/Chain of Custody Record.  Sample Delivery Group
(SDG) is a group of samples within a Case, received over a period of 7 days or
less with the same laboratory turnaround and not exceeding 20 samples
[excluding Performance Evaluation  (PE) samples].  Data for all samples in the
SDG are due concurrently.  The date of delivery of the SDG or any samples
within the SDG is the date that the last sample in the SDG is received.  See
Exhibit A for further description.

      2DELIVERABLES ARE TO BE REPORTED TOTAL AND COMPLETE.   Concurrent
delivery is required.  Delivery shall be made such that all designated
recipients receive the item on the same calendar day.  This includes
resubmission of both the hardcopy and electronic deliverable.  The date of
delivery of the SDG, or any sample within the SDG, is the date all samples
have been delivered.   If the deliverables are  due on a Saturday,  Sunday,  or
Federal holiday, then they shall be delivered on the next business day.
Deliverables received after this time will be considered late.

      'Complete SDG File (CSF) will contain the original Sample Data Package
plus all of the original documents described in Exhibit B, Section 2.6.

      4If required at the time of sample scheduling,  the Contractor shall
provide Preliminary Results, consisting of all Form Is  (see Exhibit B,
Section 2.9).  Facsimile or electronic transmittal is required as requested by
the Region.  Electronic transmittals shall be transmitted as WordPerfect, MS
Word, PDF, or other USEPA-approved formats.  The Contractor will be notified
of the format, fax numbers, or email address(es) at the time of sample
scheduling.  Sample Traffic Reports/Chain of Custody Records and SDG Cover
Sheets shall be submitted with the Preliminary Results.  The Contractor shall
document all communication in a telephone log.

Preliminary Results Delivery Schedule:

If a sample requiring Preliminary Results arrives before 5 p.m., the
Preliminary Results are due within the required turnaround time.  If a sample
requiring Preliminary Results is received after 5 p.m., the Preliminary
Results are due within the required turnaround time beginning at 8 a.m. the
following day.

      5Also required in each Sample Data Package.

      6See Exhibit E for description.  Time is cited in calendar days.
                                       B-7                                ILM05.2

-------
Exhibit B — Section 1
Contract Reports/Deliverables Distribution  (Con't)

Footnotes  (Con't):

      7The Contractor shall deliver both hardcopy and electronic  (i.e.,
diskette) copies of the Standard Operating  Procedures  (SOPs) and  Quality
Assurance Management Plan  (QAP).

1.2   Distribution

      The  following addresses correspond to the  "Distribution"  column  in
      Exhibit B, Section 1.1, Table  1.

      SMO:         USEPA Contract  Laboratory Program .(CLP)
                   Sample Management Office (SMO) x
                   2000  Edmund Halley Drive
                   Reston,  VA 20191-3400

      Region:      USEPA REGIONS:  SMO will provide the  Contractor  with  the list
                   of addressees for data delivery for  the 10 USEPA Regions.
                   SMO will provide the Contractor with updated Regional
                   address/name lists as necessary throughout the  period of the
                   contract and identify other client recipients on a case-by-
                   case  basis.

      USEPA Regional  CLP Project  Officer (CLP PO):

                   SMO will provide the Contractor with the list of addresses
                   for the USEPA Regional CLP POs.  SMO will provide the
                   Contractor with updated name/address lists as necessary
                   throughout the period of the contract.

      QATS:        USEPA Contract  Laboratory Program (CLP)
                   Quality  Assurance Technical Support  (QATS)  Laboratory2
                   2700  Chandler Avenue, Building C
                   Las Vegas, NV 89120
                   Attn: Data Audit Staff

      In addition,  the  mailing and delivery addresses  for the  USEPA AOC
      Inorganic Program Manager (AOC PM)  are:

      Mailing Address:  USEPA OERR Analytical Operations/
                         Data Quality Center
                         Ariel Rios Building  (5204G)
                         1200 Pennsylvania Avenue, N.W.
                         Washington, DC 20460
                         Attn: CLP Inorganic  Program Manager

      Fed-Ex/Overnight  USEPA OERR Analytical Operations/
        Delivery:         Data Quality Center
                         1235 Jefferson Davis Highway
                         Crystal Gateway  I, 12th  Floor
                         Arlington, VA  22202
                         Attn: CLP Inorganic  Program Manager
       1The  SMO  is  a  Contractor-operated  facility  operating under  the  SMO
 contract awarded and administered by USEPA.

       2The  QATS laboratory  is  a  Contractor-operated  facility operating under
 the QATS contract awarded and administered by USEPA.

 ILM05.2                               B-8

-------
                                                         Exhibit B  —  Section 2
                          Reporting Requirements and Order  of Data  Deliverables

2.0   REPORTING REQUIREMENTS AND ORDER OF DATA  DELIVERABLES

2.1   Introduction

      The Contractor shall provide  reports and  other deliverables as specified
      in Exhibit B, Section 1.1.  The required  content and form of  each
      deliverable is described in this exhibit.  All reports and documentation
      shall be:

      ;I!  Legible;

      111  Clearly labeled  and completed in accordance with instructions in this
         exhibit;

      :*'  Arranged in the  order specified in this section;

      ="  Paginated sequentially according to  instructions in this exhibit; and

      ="  Double-sided.

      NOTE: Complete Sample Delivery Group  (SDG) Files  (CSFs) need  not be
      double-sided.   (The CSF is composed of  original documents.)   However,
      Sample Data Packages delivered to the USEPA Contract Laboratory  Program
      (CLP) Sample Management Office  (SMO) and  the  Region,  [and USEPA
      designated recipients, e.g.,  Quality Assurance Technical Support (QATS),
      upon written request] must be double-sided.

2.1.1    The Contractor shall use EPA Case numbers, SDG numbers, and EPA
         sample numbers to identify samples received under this contract, both
         verbally and  in  reports and correspondence.  The contract  number
         shall be specified in all  correspondence.

2.1.2    Section 4 of  this exhibit  contains the required Data Reporting Forms
         in Agency-specified format.  Section 3 of  this Exhibit contains
         instructions  to  the Contractor for properly completing all data
         reporting forms  to provide USEPA with  all  required  data.   Data
      •   elements and  field descriptors for reporting data in computer-
         readable format  are contained in Exhibit H.

2.2   Resubmission of  Data

      If submitted documentation does not conform to the above criteria, the
      Contractor is required to resubmit such documentation with
      deficiency(ies)  corrected within 4 business days, at no additional cost
      to USEPA.
Ť
2.2.1    Whenever the  Contractor is required  to submit or resubmit  data as a
         result of an  on-site laboratory evaluation, through the USEPA
         Regional CLP  Project Officer  (CLP PO)  action, or through a Regional
         data reviewer's  request,  the data shall be clearly marked  as
         "Additional Data" and shall be sent  to both contractual data
         recipients  (SMO  and Region) and to USEPA's designated recipient
          (e.g., QATS)  when a written request  for the Sample  Data Package has
         been made.  A cover letter shall be  included which describes  what
         data is being delivered,  to which USEPA Case(s) the data pertains,
         and who requested the data .

2.2.2    Whenever the  Contractor is required  to submit or resubmit  data as a
         result of Contract Compliance Screening  (CCS) review by SMO,  the data
         shall be sent to the two  contractual data  recipients  (SMO  and Region)
         and to USEPA's designated  recipient  (e.g., QATS) when a written
         request for the  Sample Data Package  has been made.  In all instances,
         the Contractor shall include a color-coded cover sheet  (Laboratory

                                      B-9                                ILM05.2

-------
Exhibit B — Section 2
Reporting Requirements and Order of Data Deliverables (Con't)

         Response to Results of Contract Compliance Screening) provided by
         SMO.  Electronic deliverables shall be submitted or resubmitted to
         SMO and the Region.  Revised DC-1 and DC-2 forms shall be resubmitted
         to SMO and the Region.

2.3   Quality Assurance  (QA) Management Plan and Standard Operating Procedures
       (SOPs)

      The Contractor shall adhere to the requirements in Exhibits E and F.

2.4   Sample Traffic Reports/Chain of Custody Records

      Each sample received by the Contractor will be labeled with an EPA
      sample number and will be accompanied by a Sample Traffic Report/Chain
      of Custody Record bearing the sample number and descriptive information
      regarding the sample.  The current CLP Traffic Report  is the "Inorganic
      Traffic Report & Chain of Custody Record".  The CLP Traffic Report/Chain
      of Custody Record is one form divided into two sections: the Traffic
      Report section which consists of everything above the  Chain of Custody
      Record section, and the bottom section which is the Chain of Custody
      Record.  The Contractor shall complete the CLP Traffic Report/Chain of
      Custody Record  (marked "Lab Copy for Return to SMO"),  recording the date
      of sample receipt, verifying the number of samples, and signing the CLP
      Traffic Report/Chain of Custody Record.

      Upon receipt, the Contractor shall sign for receipt of samples.. The
      laboratory signature box is located at the bottom of the CLP Traffic
      Report/Chain of Custody Record in the Chain of Custody Record section.
      The laboratory sample custodian or designated recipient opening and
      verifying the contents of the cooler shall then verify receipt of all
      samples identified within the CLP Traffic Report section and sign and
      date the signature box located in the upper half of the CLP Traffic
      Report/Chain of Custody Record.  If a non-CLP Traffic  Report/Chain of
      Custody Record is  submitted with the samples, for example a Regional
      Traffic Report/Chain  of Custody Record, then the Contractor shall  (1)
       sign and date receipt of the samples to maintain the chain-of-custody
       and  (2) the  sample custodian or designated recipient shall sign and date
       the Traffic  Report/Chain of Custody Record to verify sample information.

       The Contractor  shall  also enter the Sample Delivery Group  (SDG) number.
       Case number, and  the  laboratory contract number on the CLP Traffic
       Report/Chain of Custody Record, in the appropriate boxes.  The EPA
       sample  number of  the  first  sample received in the SDG  is the SDG number.
       When  several samples  are received together in the first SDG shipment,
       the  SDG number  shall  be the  lowest sample number  (considering both alpha
       and numeric  designations) in  the  first group of samples received under
       the  SDG.  Under no circumstances  should any SDG number be replicated
       within  a  Case.   If necessary,  select an alternative sample number  for
       the  SDG number.   The  SDG number is also reported on all data reporting
       forms  (see  Exhibit B,  Section  3 -  Form  Instructions).   If the laboratory
       is requested to  transfer  samples  to another facility,  the Contractor
       shall  date  and  enter  the name  of  the  facility  to where the samples will
       be transferred  on the CLP Traffic  Report/Chain of Custody Record.

 2.4.1    The  Contractor shall  submit  Traffic Reports/Chain of Custody Records
          in SDG sets  (i.e..  Traffic Reports/Chain of Custody Records  for all
          samples  in an SDG shall  be clipped together), with  an  SDG Coyer Sheet
          attached.   The SDG Cover Sheet  shall  contain  the  following  items:
 ILM05.2                              B-10

-------
                                                         Exhibit B — Section 2
                 Reporting Requirements and Order of Data  Deliverables  (Con't)

         •*   Laboratory name;

         •*=   Contract number;

         ^   Sample analysis price  (full sample price from the contract);

         ť•   Case number; and

         *   List of EPA sample numbers of all samples in  the SDG, identifying
             the first and last  samples received,  and  their Laboratory Receipt
             Dates  (LRDs).

         NOTE: When more than one sample is received in the first or last SDG
         shipment, the "first" sample received would be the sample with the
         lowest sample number  (considering both alpha and  numeric
         designations); the "last"  sample received would be the sample with
         the highest sample number  (considering both alpha and numeric
         designations).

2.4.2    EPA field sample numbers are six digits in length and continuous
         (without spaces or hyphens).   If the Contractor receives sample
         numbers of any other length, the Contractor shall contact SMO
         immediately.  The original Sample Traffic Report/Chain of Custody
         Record page marked "Lab Copy for Return to SMO",  with laboratory
         receipt information and signed with original Contractor signature,
         shall be submitted for each sample in the SDG.

2.4.3  -  If samples are received at the laboratory with multi-sample Traffic
         Reports/Chain of Custody Records, all the samples on one multi-sample
       . . Traffic Report/Chain of Custody Record may not necessarily be in the
         same SDG.  In this instance, the Contractor shall make the
         appropriate number of photocopies of the Traffic  Report/Chain of
       1  Custody Record, and submit one copy with each SDG Cover Sheet.

2.5   Sample Data Package

      The Sample Data Package shall include data for analysis of all samples
      in one SDG, including field and analytical samples,  blanks, spikes,
      duplicates, and Laboratory Control Samples (LCSs).   The Sample Data
      Package shall be complete before  submission,  and shall be consecutively
      paginated  (starting with page number one and ending  with the number of
      all pages in the package).  The Sample Data Package  shall include the
      following:

2.5.1    Cover Documentation

2.5.1.1     Cover Page for the inorganic analyses Data Package shall include:
            laboratory name; laboratory code; contract number; Case number;
            SDG number; Non-Routine Analytical Service  (NRAS) number  (if
            appropriate); EPA sample numbers in alphanumeric order showing EPA
            sample numbers cross-referenced with laboratory Sample ID numbers;
            and completion of the questions on use of background and
            interelement corrections for the samples.

2.5.1.1.1      The Cover Page shall contain the following  statement, verbatim:
               "I certify that this Sample Data Package is in compliance with
               the terms and conditions of the contract, both technically and
               for completeness, for other than the conditions detailed above.
               Release of the data  contained in this hardcopy Sample Data
               Package and in the computer-readable data submitted on diskette
                (or via an alternate means of electronic transmission, if
               approved in advance  by USEPA) has been authorized by the
               Laboratory Manager or the Manager's designee, as verified by

                                      B-ll                               ILM05.2

-------
Exhibit B — Section 2
Reporting Requirements and Order of Data Deliverables  (Con't)

                the  following signature."   This  statement  shall  be  directly
                followed  by  the  signature  of the Laboratory Manager or designee
                with typed lines containing the  signer's name and title,  and
                the  date  of  signature.

2.5.1.2     SDG Narrative.   This document  shall be clearly labeled "SDG
            Narrative" and  shall contain:  laboratory  name. Case number,  SDG
            number,  contract number,  and  detailed documentation of any Quality
            Control (QC), sample,  shipment,  and/or analytical problems
            encountered  in  processing the samples reported in the  Sample Data
            Package.  The Contractor  shall include any technical and
            administrative  problems encountered and the resolution or
            corrective actions  taken.   This includes  documenting the
            alternative  technique  used to determine cooler temperature if a
            temperature  indicator  bottle  is not present in the  cooler.  The
            Contractor shall also  provide, in the SDG Narrative, sufficient
            information,  including equations or curves (at least one equation
            or curve per method),  to  allow the  recalculation of sample results
            from raw instrument output.  The Contractor shall also include a
            discussion of any flexibility Statement of Work  (SOW)
            modification.  This includes  attaching a  copy of the USEPA
            approved modification  form to the SDG Narrative. Additionally the
            Contractor shall also  identify and explain any differences which
            exist between the Form Is and supporting  documentation provided in
            the data package and those previously provided as Preliminary
            Results.

2.5.1.3     Sample  Log-In Sheet [Form DC-1]

2.5.1.4     Full  Inorganics Complete  SDG  File (CSF) Inventory Sheet [Form DC-
            2]

2.5.2     Sample Data

          Sample data shall  be submitted with the inorganic analysis data
          reporting forms for all samples  in the SDG.   Data should  be arranged
          in increasing alphanumeric EPA sample number order, followed by the
          QC analyses data,  quarterly and annual verification of method and
          instrument parameters  forms, raw data, and copies of the  digestion
          and  distillation logs.

2.5.2.1     Inorganic Analysis  Data Sheet [Form IA-IN and Form  IB-IN].
             Tabulated analytical results of the requested analytes shall be
             included.  The  validation and release of  these results is
             authorized by a specific signed statement on  the Cover Page.  In
             the event that  the laboratory cannot verify  all  data reported for
             each sample, the Laboratory Manager shall provide a detailed
             description  of  the problems associated with  the  sample(s) in the
             SDG Narrative.

2.5.2.1.1      Appropriate  concentration units shall  be  specified  and entered
                on Forms IA-IN and IB-IN.  The quantitative values  shall  be
                reported in  units of micrograms per Liter  (UG/L) for water
                samples and milligrams per kilogram (MG/KG) for solid samples.
                (No other units are acceptable.)  Results  for solid samples
                shall be reported on a dry weight basis.   Analytical results
                shall be reported to two significant figures  if the result
                value is less than 10 and  to three significant figures if the
                value is greater than or equal  to 10.   Results for percent
                solids shall be reported to one decimal place.  The preceding
                discussion concerning significant numbers  applies to Forms IA-
                IN, IB-IN, and IX-IN only.  For other forms,  follow the
 ILM05.2                               B-12

-------
                                                         Exhibit B —  Section  2
                 Reporting Requirements  and Order  of  Data  Deliverables (Con't)

               instructions  specific to  those  forms as  discussed in  this
               exhibit.

2.5.2.2     Quality Control  (QC) Data

2.5.2.2.1      The QC summary for inorganic analysis  shall contain the forms
               listed below.

               NOTE: If more than one  form is  necessary, duplicate forms  must
               be arranged in chronological order.

2.5.2.2.1.1       Initial and Continuing Calibration  Verification  [Form IIA-
                  IN]

2.5.2.2.1.2       CRQL Check Standard  [Form IIB-IN]

2.5.2.2.1.3       Blanks  [Form  III-IN]

2.5.2.2.1.4       ICP-AES Interference Check Sample [Form  IVA-IN]

2.5.2.2.1.5       ICP-MS  Interference Check Sample  [Form IVB-IN]

2.5.2.2.1.6       Matrix  Spike  Sample Recovery [Form  VA-IN]

2.5.2.2.1.7       Post-Digestion Spike Sample  Recovery  [Form VB-IN]

2.5.2.2.1.8       Duplicates [Form VI-IN]

2.5.2.2.1.9       Laboratory Control Sample [Form VII-IN]

2.5.2.2.1.10      ICP-AES and ICP-MS Serial Dilutions [Form VIII-IN]

2.5.2.2.1.11      Method  Detection Limits  (Annually)  [Form IX-IN]

2.5.2.2.1.12      ICP-AES Interelement Correction Factors  (Quarterly)
      -             [Form XA-IN]

2.5.2.2.1.13      ICP-AES Interelement Correction Factors  (Quarterly)
                   [Form XB-IN]

2.5.2.2.1.14      ICP-AES and ICP-MS Linear Ranges  (Quarterly)  [Form XI-IN]

2.5.2.2.1.15      Preparation Log  [Form  XII-IN]

2.5.2.2.1.16      Analysis Run  Log  [Form XIII-IN]

2.5.2.2.1.17      ICP-MS  Tune  [Form XIV-IN]

2.5.2.2.1.18      ICP-MS  Internal Standards Relative  Intensity  Summary [Form
                  XV-IN]

2.5.2.3     Raw Data

            For each reported value, the Contractor shall  include  in the
            Sample  Data Package all raw  data 'used  to  obtain that value.   This
            applies to all required QA/QC measurements,  instrument
            standardization, as well as  all sample  analysis results.  This
            statement does not  apply to  the quarterly and  annual verification
            of method and instrument parameters submitted  as a  part  of each
            Sample  Data Package.  When analysis of  the  ICP-AES  or  ICP-MS
            target  analytes  listed  in  Exhibit  C of  this SOW  (or any  subset or
            additional analytes) is requested, the  raw  data shall  include, for
            all samples,  not only the  results  for  the requested analyte(s),

                                      B-13                               ILM05.2

-------
Exhibit B — Section 2
Reporting Requirements and Order of Data Deliverables  (Con't)

            but also those  for  all  the  interferents  (Exhibit  D/ICP-AES,  Table
            1, or Exhibit D/ICP-MS, Section  7.2.4.4.1,  as  appropriate).   The
            raw data shall  also contain the  results  of  any other analyte(s)
            which have been  determined  to  interfere  with the  requested
            analytes(s).

2.5.2.3.1      Raw data shall contain all  instrument readouts and data
               pertinent  to the reconstruction of  the  analysis and results
                (e.g., Batch Sheets) used for the sample results.  Each
               exposure or  instrumental reading shall  be provided, including
               those readouts  that  may  fall  below  the  Method Detection Limit
                (MDL).   Raw  data shall not  be corrected for dilutions or volume
               adjustments.  All  Atomic Absorption (AA),  Inductively Coupled
               Plasma - Atomic  Emission Spectrometer (ICP-AES),  and
               Inductively  Coupled  Plasma  -  Mass Spectrometry (ICP-MS)
               instruments  shall  provide a legible hardcopy of the direct
               real-time  instrument readout  (i.e., strip charts, printer
               tapes, etc.)  or  a  printout  of the unedited  instrument data
               output  file.  A photocopy of  the instrument's direct sequential
               readout  shall be included.   A hardcopy  of the instrument's
               direct  readout  shall be  included for cyanide if the
               instrumentation has  the  capability.

2.5.2.3.2      The  order  of raw data  in the  Sample Data Package for inorganic
               analyses shall  be: ICP-AES, Graphite Furnace Atomic Absorption
                (GFAA),  ICP-MS,  Mercury, and Cyanide.  All  raw data shall
               include  concentration units for ICP,  and absorbances or
               concentration units  for  Mercury and Cyanide.

2.5.2.3.3      The  ICP-MS raw  data  shall also contain  the  turbidity
               measurements results [in Nephelolometric Turbidity Units  (NTU)]
                for the  field samples.

2.5.2.3.4      Corrections  to  the laboratory data  reporting forms and raw data
                shall be made by drawing single lines through the errors and
                entering the correct information.   Information shall not be
                obliterated or rendered unreadable.  Corrections and additions
                to information shall be  signed  (or initialed) and dated.

2.5.2.3.5      Raw data shall  be  labeled with EPA  sample numbers and
                appropriate codes, shown in Exhibit B,  Table 2 - Codes for
                Labeling Data,  following, to unequivocally identify:

                lx:   Calibration standards,  including source and preparation
                    date.   Standard preparation logbooks can be  submitted if
                    they contain this information;

                ^   Initial and Continuing Calibration Blanks  (ICBs/CCBs) and
                    Preparation Blanks  (PBs);

                "=   Initial and Continuing Calibration Verification  (ICV/CCV)
                    standards,   Interference Check  Samples  (ICSs),  serial
                    dilution samples, Contract Required Quantitation Limit
                     (CRQL) Check Standard  (CRI), LCS, and post digestion  spike;

                00   Diluted and undiluted  samples  (by EPA sample number)  and
                    all weights, dilutions, and volumes used  to  obtain the
                    reported values (if the volumes, weights, and  dilutions are
                    consistent  for all  samples  in  a  given SDG, a general
                    statement outlining these parameters is sufficient);

                <*   Duplicates;
 ILM05.2                               B-14

-------
                                                        Exhibit B — Section 2
                Reporting Requirements and Order of  Data  Deliverables  (Con't)

                  Spikes  (indicating  standard  solutions used,  final  spike
                  concentrations,  and volumes  involved).   If  spike
                  information  (source, concentration,  volume)  is  consistent
                  for  a given  SDG,  a  general statement outlining  these
                  parameters is  sufficient;

                  Instrument used,  any instrument  adjustments, data
                  corrections  or other apparent anomalies on  the  measurement
                  record, including all data voided  or data not used to
                  obtain reported  values  and a brief written  explanation;  and

                  Time and  date  of each analysis.  Instrument run logs can
                  also be submitted if they contain  time  and  date of
                  analysis.  If  the instrument does  not automatically provide
                  times of  analysis,  these shall be  manually  entered on  all
                  raw  data  (e.g.,  ICV/CCV, blanks, and the CRQL Check
                  Standard).
                                   Table 2
                          Codes for Labeling Data 1-2-3
Sample
Sample Not Part of the SDG
Duplicate
Matrix Spike
Serial Dilution
Analytical Spike/Post
Digestion/Distillation Spike

Instrument Calibration Standards:
        ICP
        Atomic Absorption and Cyanide
        Initial Calibration Verification
        Initial Calibration Blank
Continuing Calibration Verification
Continuing Calibration Blank
Interference Check Samples:
        Solution A
        Solution AB
CRQL Check Standard
Laboratory Control Samples:
        Aqueous  (Water)
        Solid  (Soil/Sediment)
Preparation Blank (Water)
Preparation Blank (Soil)
Linear Range Analysis Standard
Baseline Correction
Reslope
Cyanide Mid-Range Standard
ICP-MS Tune Check
                    XXXXXX
                    ZZZZZZ
                   XXXXXXD
                   xxxxxxs
                   XXXXXXL
                   XXXXXXA
S or SO for blank standard
           SO,  S10,...etc.
                     ICV##
                     ICB##
                     CCVff
                     CCBff

                    ICSAff
                   ICSABtt
                     CRI##

                    LCSW#f
                    LCSSff
                     PBW##
                     PBSf#
                     LRSf#
                BASELINEff
                 RESLOPEf*
                MIDRANGE##
                    TUNE##
                                     B-15
                    ILM05.2

-------
Exhibit B — Section 2
Reporting Requirements and Order of Data Deliverables  (Con't)

      lThe numeric suffix that follows the "S" suffix for the standards
indicates the true value of the concentration of the standard in ug/L.

      2ICP-AES and ICP-MS calibration standards usually consist of several
analytes at different concentrations.  Therefore, no numeric suffix can follow
the ICP calibration standards unless all the analytes  in the standard are
prepared at the same concentrations.  For instance, the blank for  ICP shall be
formatted "SO".

      3The EPA sample number shall be unique for each ICV,  ICB, CCV, CCB,
ICSA, ICSAB, CRI, LCSW, LCSS, PBW, PBS, LRS, BASELINE, RESLOPE, MIDRANGE, and
TUNE within an analysis or preparation method, within  an SDG.  The Contractor
shall achieve this by replacing the two-character terminator  (##)  of the
identifier with one or two characters, numbers, or a combination of both.

2.5.2.4     Digestion and Distillation Logs.  The following  logs shall be
            submitted as appropriate  for  each preparation  procedure:  digestion
            logs  for ICP-AES,  ICP-MS, mercury preparations,  and cyanide.
            These logs  shall  include:  (1)  date;  (2)  sample weights and
            volumes, with  initial  sample  weight/volume and final volume
            clearly  indicated;  (3)  sufficient information  to unequivocally
            identify which  QC samples (i.e.,  LCS,  PB)  correspond to each  batch
            digested;  (4)  comments describing any significant sample changes
            or  reactions which occur  during preparation shall be entered  in
            the log  and noted in  the  SDG  Narrative;  (5)  indication of pH  less
            than  2 or  greater than 12,  as applicable;  and  (6)  identification
            of  the sample  preparer(s)  [signature(s)].

2.5.3     A  copy of the  Sample Traffic Reports/Chain  of Custody Records
          submitted in  Exhibit B,  Section  2.4,  for all  of the samples in the
          SDG.   The Traffic  Reports/Chain  of Custody  Records  shall  be arranged
          in increasing  EPA sample  number  order,  considering  both alpha and
          numeric  designations.  A  legible photocopy  of the SDG Cover Sheet
          shall  also  be  submitted.

2.6    Complete  SDG File (CSF)

       As specified in  the  Delivery Schedule,  one CSF (including the original
       Sample  Data Package)  shall be delivered to the Region  concurrently  with
       the delivery of  a copy  of the Sample Data Package to SMO.   Delivery to
       USEPA's designated recipient (e.g.,  QATS)  is only required upon written
       request.

2.6.1     The  CSF  shall  contain all original documents  where  possible.   No
          photocopies  of original  documents shall be  placed in the  CSF unless
          the  original  data was initially written in  a bound  notebook,
          maintained  by the Contractor,  or the originals were previously
          submitted to  USEPA with another Case/SDG in accordance with the
          requirements  described in Exhibit F.  The CSF shall contain all
          original documents and be numbered according to the specifications in
          Exhibit B,  Sections  3 and 4, and Form DC-2.

2.6.2     The  CSF shall  consist of  the following original documents in addition
          to the documents  in  the Sample Data Package.

          NOTE:  All Case-related documentation may be used  or admitted as
          evidence in subsequent legal proceedings.   Any other Case-specific
          documents generated after the CSF is sent to USEPA, as well as copies
          that are altered in any fashion, are also deliverables to USEPA.
          Send the original to the Region and a copy to SMO.   Send to USEPA's
          designated recipient (e.g.,  QATS) only upon written request.
 ILM05.2                               B-16

-------
                                                         Exhibit B — Section  2
                 Reporting Requirements  and  Order  of  Data  Deliverables  (Con't)

2.6.2.1     Original Sample Data Package

2.6.2.2     A completed and signed Full  Inorganics Complete SDG File  (CSF)
            Inventory Sheet  [Form DC-2]

2.6.2.3     All original shipping documents, including, but not limited to,
            the following documents:

            '*=  USEPA Sample Traffic Reports/Chain of Custody  Records

            "=  Airbills  (if an airbill  is not  received, include a hardcopy
                receipt requested from the shipping company or a printout of
                the shipping company's electronic  tracking information);  and

            'Ť=  Sample Tags  (if present)  sealed in plastic bags.

2.6.2.4     All original receiving documents, including, but not limited  to,
            the following documents:

            •"  Form DC-1;

            ^  Other receiving forms or copies of receiving logbooks;  and

            Ť•  SDG Cover Sheet.

2.6.2.5     All original laboratory records  of  sample transfer, preparation,
            and analysis, including, but not limited  to, the following
            documents:

            =*=  Original preparation and analysis  forms  or copies of
                preparation  and analysis logbook pages;  and

            ^  Internal sample and sample digestate  and distillate transfer
                Chain of Custody Records.

2.6.2.6     All other original SDG-specific  documents in the possession of  the
            laboratory, including, but not limited to, the following
            documents:

            ="  Telephone contact logs;

            -*=  Copies of personal logbook pages;

            'x:  All handwritten SDG-specific notes; and

            :x:  Any other SDG-specific documents not  covered by the above.

2.6.3    If the Contractor does submit SDG-specific documents  to USEPA after
         submission of the CSF, the documents shall be numbered as an addendum
         to the CSF and a revised Form DC-2  shall  be  submitted; or the
         documents shall be numbered as  a new CSF  and a new Form DC-2 shall be
         submitted to the Region only.

2.6.4    The Contractor shall retain a legible  electronic  (PDF) or hard copy
         of the CSF for 365 days after submission  of  the reconciled data
         package.  After this time, the  Contractor may dispose of the package.

2.7   Data in Computer-Readable Format

      The Contractor shall provide a computer-readable copy for all samples in
      the SDG, as specified in Exhibit H, and delivered as specified in
      Exhibit B, Section 1.1.  Computer-readable data deliverables shall  be
      submitted on DOS formatted 3.5-inch high  density 1.44 MB diskette (s)  (or

                                      B-17                               ILM05.2

-------
Exhibit B — Section 2
Reporting Requirements and Order of Data Deliverables  (Con't)

      via an alternate means of electronic transmission, if  approved in
      advance by USEPA).

2.7.1    When submitted, diskette(s) shall be packaged and shipped in such a
         manner that the diskette(s) cannot be bent or folded and will not be
         exposed to extreme heat/cold or any type of electromagnetic
         radiation.  The diskette(s) shall be included in the same shipment as
         the hardcopy data, and, at a minimum, be enclosed in a  diskette
         mailer.

2.7.2    The data  shall be recorded in the file  format and adhere to the  file,
         record, and field specifications listed in Exhibit  H, "Data
         Dictionary and Format for Data Deliverables in  Computer-Readable
         Format".

2.8   Results of the Intercomparison and Performance Evaluation  (PE) Sample
      Analyses

      Tabulation of analytical results for intercomparison/PE sample analyses
      includes all requirements specified in Exhibit B,  Sections 2.5 and  2.7.

2.9   Preliminary  Results

      The Form Is  data results (including all appropriate qualifiers and
      flags) shall be  submitted for all samples  in one SDG of a  Case.  Sample
      analysis shall follow all requirements stipulated  in Exhibit D.  The
      Contractor shall clearly identify the Preliminary  Results  by labeling
      each  Form  I  as "Preliminary Results" under the form title  (e.g., under
      Inorganic Analysis  Data  Sheet).  The Contractor  shall  also include  a
      disclaimer in the "Comments"  field on all  Form Is  stating  that the  "Data
      results contained on this Form  I are for scanning  purposes only, and may
      not have been validated  for CLP criteria." Sample Traffic Reports/Chain
      of Custody Records  and  SDG  Cover Sheets shall be submitted with the
      Preliminary  Results.

2.9.1    The Contractor shall  submit'the Cover Page following the
          specifications  in Exhibit  B, Sections 2.5.1 and 3.4.1.  The Cover
          Page  shall be clearly labeled to  indicate  that  the  data being
          reported  are  Preliminary Results.  The  Cover  Page  shall contain  the
          following statement,  verbatim : "I certify that these Preliminary
          Results are  in compliance  with  the terms and  conditions of  the
          contract, both technically and  for completeness,  for other  than  the
          conditions detailed above.   Release  of  the data contained in this
          hardcopy  data package has  been  authorized  by  the  Laboratory Manager
          or the  Manager's designee,  as verified  by  the following signature."
          This  statement shall be  directly  followed  by  the  signature  of  the
          Laboratory Manager  or designee  with  typed  lines containing  the
          signer's  name and title,  and the  date of  signature.

2.10 Quarterly  Verification  of Linear Ranges and Interelement Correction
       Factors  and  Annual  Verification of MDLs

       The Contractor  shall perform  and  report  quarterly verification of
       instrument linear range and annual verification  of MDLs by the methods
       specified  in Exhibit D for  each instrument used  under this contract.
       The Contractor  shall also perform and report  quarterly ICP-AES
       interelement correction factors (including method of determination),
       wavelengths  used,  and integration  times.   Forms  reporting  results  for
       quarterly  and annual verification  of method and  instrument parameters
       for the  current quarter and year  shall  be  submitted in each  Sample  Data
       Package,  using  Inorganic Forms IX, XA,  XB, and  XI.  Submission of  the
       quarterly  and annual verification  of method and  instrument parameters
       shall include  the raw data  used to determine  the values  reported.

 ILM05.2                              B-18

-------
                                                         Exhibit B — Section 2
                 Reporting Requirements and Order of Data Deliverables  (Con't)

2.11  Electronic Instrument Data

      The Contractor shall adhere to the requirements in Exhibit E.

2.12  Corrective Action Procedures

      If the Contractor fails to adhere to the requirements detailed in this
      SOW, the Contractor will be in noncompliance with the contract and may
      be subjected to sanctions as described in the contract.
                                      B-19                               ILM05.2

-------
Exhibit B — Section 3
Form Instructions

3.0   FORM INSTRUCTIONS

3.1   Introduction

      This section contains  specific  instructions  for the  completion  of all
      required  Inorganic  Data  Reporting  Forms.

3.2   General Information

      Values shall be  reported on  the hardcopy  forms according  to  the
      respective  form  instructions in this  section.  Each  form  submitted  shall
      be  filled out  completely for all analytes before  proceeding  to  the  next
      form of the same type.   Do not  submit multiple forms if the  information
      on  those  forms can  be  submitted on one form.

3.2.1     The data reporting  forms  discussed in  Exhibit  B,  Section  3.4, and
          presented in  Exhibit  B, Section 4.0, have been designed in
          conjunction with the  computer-readable data formats specified in
          Exhibit  H,  "Data Dictionary  and Format for Data Deliverables in
          Computer-Readable Format".  The specific  length of each variable for
          computer-readable data transmission purposes  is given  in Exhibit H.
          Information entered on these forms shall   not  exceed the  size of the
          field  given on the  form,  including such laboratory-generated items  as
          "Lab Name"  and "Lab Sample ID".

          NOTE:  On the  hardcopy forms, the space provided for entries  is
          greater  in  some instances than  the length prescribed for the variable
          as  written  to the electronic deliverable  (see  Exhibit  H). Greater
          space  is provided on the  hardcopy forms for  the sake of visual
          clarity.

3.2.2     All characters which  appear  on  the data reporting forms presented in
          the contract  shall be reproduced by the Contractor when submitting
          data,  and the format of the  forms submitted  shall be identical  to
          that  shown  in the contract.   No information may be added, deleted,  or
          moved  from  its specified position without prior written approval of
          the USEPA Regional Contract Laboratory Program Project Officer   (CLP
          PO) or the  USEPA OERR Analytical Operations/Data Quality Center (AOC)
          Inorganic Program Manager (AOC  PM).  The  names of various fields and
          analytes (i.e.,  "Lab Code",  "Aluminum")  shall appear as they do on
          the forms in  the contract, including the  options  specified in the
          form   (i.e., "Matrix  (soil/water):" shall  appear,  not just "Matrix").

3.2.3     Alphabetic  entries made onto the forms by the  Contractor shall  be in
          ALL UPPERCASE letters (i.e., "LOW", not  "Low" or "low").  If an entry
          does  not fill the entire blank  space provided on the form, null
          characters  shall be used to remove the remaining underscores that
          comprise the  blank line  (see Exhibit H for additional instructions).
          However, do not remove the underscores or vertical bar characters
          that  delineate "boxes" on the forms.

 3.3   Header Information

       Six pieces of information are common to the  header sections of  each data
       reporting form.   These are:  Laboratory Name, Contract, Laboratory Code,
       Case number,  Non-Routine Analytical Services (NRAS)  number, and Sample
       Delivery Group  (SDG) number.  Except as noted for NRAS number,  this
       information shall be entered on every form and shall match on all  forms.

 3.3.1     Laboratory Name.  The "Lab Name" shall be the name chosen by the
          Contractor to identify the laboratory.  It may not exceed 25
          characters.
 ILM05.2                               B-20

-------
                                                         Exhibit B — Section 3
                                                      Form Instructions  (Con't)

3.3.2    Contract.  The "Contract" is the number of the USEPA contract under
         which the analyses were performed.

3.3.3    Laboratory Code.  The "Lab Code" is an alphabetic abbreviation of up
         to six characters, assigned by USEPA, to identify the laboratory and
         aid in data processing.  This laboratory code will be assigned by
         USEPA at the time a contract is awarded.  The laboratory code shall
         not be modified by the Contractor, except at the direction of USEPA.
         If a change of name or ownership occurs at the laboratory, the
         laboratory code will remain the same until the Contractor is directed
         by USEPA to use another laboratory code.

3.3.4    Case Number.  The "Case No." is the SMO-assigned Case number (to five
         characters) associated with the sample, and reported on the Traffic
         Report/Chain of Custody Record.

3.3.5    NRAS Number.  The "NRAS No." is the USEPA assigned number for
         analyses performed under Non-Routine Analytical Services  (NRAS).  If
         samples are to be analyzed under NRAS only, and reported on these
         forms, then enter the NRAS number and leave the Case number blank.
         If samples are analyzed according to the Routine Analytical Services
          (RAS) protocol and have additional NRAS requirements, list both   the
         Case number and NRAS number on all forms.  If the analyses have no
         NRAS requirements, leave the "NRAS No." field blank.

3.3.6    SDG Number.  The "SDG No." is the Sample Delivery Group  (SDG) number.
         The SDG number is the EPA sample number of the first sample received
         in the SDG, except when this would cause duplication.  When several
         samples are received together in the first SDG shipment, the SDG
         number shall be the lowest sample number  (considering both alpha and
         numeric designations) in the first group of samples received under
         the SDG.  If fractions of the same field samples are scheduled under
         different turnaround times, thus creating separate SDGs containing
         the same sample numbers, a different sample number shall be utilized
         in the assignment of the SDG number for each SDG.  If a situation
         arises where there are an insufficient number of samples for
         assignment of SDG numbers, the contractor shall contact SMO for the
   *      assignment of a SDG number.

3.3.7    Sample Number.  The "EPA Sample No." appears either in the header
         information of the form or as the left column of a table summarizing
         data from a number of samples.  When an EPA sample number is entered
         in the triple-spaced box in the upper right-hand corner of a form, it
         shall be centered on the middle line of the three lines that form the
         box.

3.3.7.1     All samples, matrix spikes, post digestion/distillation spikes,
            duplicates, and serial dilutions shall be identified with an EPA
            sample number.  For samples, an EPA sample number is the unique
            identifying number given in the Traffic Report/Chain of Custody
            Record that accompanied that sample.  In order to facilitate data
            assessment, the sample suffixes listed in Exhibit B, Table 2 -
            Codes for Labeling Data, must be used.

3.3.8    Other Common Fields.  Other pieces of information are common to many
         of the data reporting forms.  These include Matrix and Level.

         :*:  For  "Matrix", enter "SOIL" for soil/sediment samples and "WATER"
             for  water samples.

             NOTE: The matrix must be spelled out.  Abbreviations such as "S"
             or "W"  shall not be used.
                                      B-21               •                ILM05.2

-------
Exhibit B — Section 3
Form Instructions                        -                                '
Cover Page

         '•"  For  "Level",  enter the determination of concentration level.
             Enter  as  "LOW"  or "MED",  not "L" or "M".

3.3.9    Rounding Rule.  For rounding off numbers to the appropriate level of
         precision, observe  the following common rules.  If the figure
         following  those to  be retained is greater than or equal to 5, round
         up; otherwise round down.  Also see  "Rounding Rules" in Exhibit G.

3.3.9.1     Before  evaluating a number for being in control or out of control
            of  a  certain limit  [other than the Contract Required Quantitation
            Limit (CRQL)], the number evaluated  shall be rounded using the
            above rounding rules  to the significance reported for that limit.
            For example, the control limit for an Initial Calibration
            Verification is  plus  or minus 10% of the true value.  Then a
            calculated percent recovery of 110.46 shall be reported on
            Form  IIA-IN as 110, which is within  the control limits of 90-110.
            On  the  other hand, a  calculated percent recovery of 110.50 shall
            be  reported on Form IIA-IN as 111, which is not within the 90-110
            percent control  limits.

            NOTE: All  results  shall be transcribed to Inorganic Forms IIA-IN
            through XV-IN  from the raw data to the specified number of decimal
            places  that are  described in Exhibits B and H.  The raw data
            result  is  to be  rounded only when the number of figures in the  raw
            data  result exceeds the maximum number of figures specified for
            that  result entry  for that form.  If there are not enough figures
            in  the  raw data  result to enter in the specified space for that
            result, then zeros shall be used  for decimal places to the
            specified  number of reporting decimals for that result for a
            specific  form.  The following examples are provided:
Raw Data Result
95.
95.
95.
95.
95.
99653
99653
99653
996
9
5
5
5
5
5
.4
.3
.2
.4
.4
(to
(to
(to
(to
(to
Specified Format Correct Entry on Form
four
three
decimal
decimal
places)
places)
two decimal places)
four
four
decimal
decimal
places)
places)
95.
95.
96.
95.
95.
9965
997
00
9960
9000
 3.4    Inorganic Forms

 3.4.1     Cover Page - [COVER PAGE]

 3.4.1.1      Purpose.  This form is  used to list all  samples  analyzed within an
             SDG and provide certain analytical information and general
             comments.  It is also the document that  is signed by the
             Laboratory Manager to authorize and release all  data and
             deliverables associated with the SDG.

 3.4.1.2      Instructions.  Complete the header information according to the
             instructions in Exhibit B, Section 3.3.   Complete the remainder of
             the form using the following instructions.

 3.4.1.2.1      For samples analyzed using this Statement of  Work (SOW), enter
                "ILM05.2" for the SOW Number.

 3.4.1.2.2      Enter an EPA sample number including  spikes and duplicates (to
                seven spaces) of every sample analyzed within the SDG.  Spikes
                shall contain an "S" suffix and duplicates a "D" suffix.  These
                sample numbers shall be listed on the form in ascending
                alphanumeric order.  Thus, if MAB123 is the lowest (considering

 ILM05.2                                B-22

-------
                                                         Exhibit B — Section 3
                                                              Form Instructions
                                                          Forms IA-IN and IB-IN

               both alpha and numeric characters) EPA  sample number within the
               SDG, it would be entered in the  first EPA sample number  field.
               Samples would be listed below  it,  in ascending sequence  -
               MAB124, MAB125, MAC111, MA1111,  MA1111D, etc.

3.4.1.2.3      A maximum of 20 field sample numbers (excluding PE samples) can
               be entered on this  form.  Submit additional Cover Pages, as
               appropriate, if the total number of samples, duplicates, and
               spikes in the SDG is greater than 22.

3.4.1.2.4      A Laboratory Sample ID (to ten spaces)  may be entered for each
               EPA sample number.  If a Laboratory Sample ID is entered, it
               shall be entered identically  (for each  EPA sample number) on
               all associated data.

3.4.1.2.5      Enter "YES" or "NO" in answer  to each of the two questions
               concerning Inductively Coupled Plasma - Atomic Emission
               Spectrometer  (ICP-AES) and Inductively  Coupled Plasma -  Mass
               Spectrometry  (ICP-MS) corrections.  Each question shall  be
               explicitly answered with a "YES" or a "NO".  The third question
               shall be answered with a "YES" or "NO"  if the answer to  the
               second question is  "YES".  It  shall be  left blank if the answer
               to the second question is "NO".

3.4.1.2.6      Under "Comments", enter any statements  relevant to the analyses
               performed under the SDG as a whole.

3.4.1.2.7      Each Cover Page shall be signed  and dated, in original,  by the
               Laboratory Manager  or the Manager's designee to authorize the
               release and verify  the contents  of all  data and deliverables
               associated with an  SDG.

3.4.2    Inorganic Analysis Data Sheet [Forms IA-IN and IB-IN]

3.4.2.1     Purpose.  These forms  are used to tabulate and report sample
            analysis results for inorganic target analytes  (see Exhibit C).

3.4.2.2     Instructions.  Complete the header  information according to the
            instructions in Exhibit B, Section  3.3.  Complete the remainder of
            the form using the following instructions.

3.4.2.2.1      "Date Received" is  the date  (formatted MM/DD/YYYY) of sample
               receipt at the laboratory, as  recorded  on the Traffic
               Report/Chain of Custody Record [i.e., the Validated Time of
               Sample Receipt  (VTSR)].

3.4.2.2.2      "% Solids" is the percent of solids on a weight-by-weight basis
               in the sample which is determined by drying the sample as
               specified in Exhibit  D - Introduction to Analytical Methods,
               Section 1.6.  Report percent solids to  one decimal place (i.e.,
               5.3%).  If the percent solids  is not required because the
               sample is fully aqueous, or is less than 1% solid, then  enter
               "0.0".

3.4.2.2.3      Enter the appropriate concentration units (UG/L for water or
               MG/KG dry weight for  soil).  Entering "MG/KG" means "mg/kg dry
               weight" on this form.

3.4.2.2.4      Under the column labeled "Concentration", enter for each
               analyte, the value  of the result [if the concentration is
               greater than or equal to the Method Detection Limit  (MDL)]
               corrected for any dilutions; or, enter  the CRQL for the

                                      B-23                              ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Forms IA-IN and IB-IN  (Con't)

               analyte,  adjusted if  necessary  and corrected for  any  dilutions,
               if  the  concentration  is  less  than  the MDL.   The concentration
               result  shall  be  reported to two significant figures if  the
               result  is less than 10 or three significant figures if  the
               value is  greater than or equal  to  10.

3.4.2.2.5      Under the columns  labeled "C",  "Q", and  "M",  enter result
               qualifiers as identified below.  If additional qualifiers are
               used, their explicit  definitions shall be included on the Cover
               Page in the "Comments" section.

               Forms IA-IN and  IB-IN include fields  for three types  of result
               qualifiers.  These qualifiers shall be completed  as follows:

3.4.2.2.5.1        C  (Concentration)  Qualifier.  Enter "J"  if the reported
                   value  was obtained from a  reading  that was less  than the
                   CRQL but greater  than or equal  to  the MDL.  If the reading
                   was  less than the  MDL, a "U" shall be entered.

                   The  MDL obtained  for  a given preparation method, analysis
                   method, and instrument shall be used  for qualification  of
                   the  results for samples associated with that  preparation
                   method, analysis  method, and instrument.  Serial dilution
                   and post-digestion spike results shall be qualified using
                   the  MDL and CRQL  values utilized for  the corresponding  field
                   sample.

                   All  three values   (i.e., the instrument reading,  CRQL, and
                   MDL) shall be converted to the  same units prior to
                   determining the appropriate C  (Concentration)  Qualifier.

                   NOTE:  The water CRQL   (in ug/L)  and the MDL obtained from
                   direct analysis (Preparation Method "NP1") for a given
                   analysis method and instrument  shall  be used to qualify the
                   results of samples and instrument QC standards that are not
                   taken through a preparation procedure [e.g.,  ICP-MS samples
                   with turbidity less  than 1 Nephelolometric Turbidity Unit
                   (NTU), ICB, CCB,  and CRI for ICP-AES].

 3.4.2.2.5.2        Q Qualifier.   Specified entries and their meanings are  as
                   follows:

                   E:  The reported value is estimated due to the presence  of
                   interference.  An explanatory note shall be included under
                   "Comments" on the Cover Page (if the problem applies to all
                   samples), or on the specific Form IA-IN or Form IB-IN (if it
                   is an  isolated problem).

                   N:  Spiked sample recovery not within control limits.

                   *:  Duplicate analysis not within control limits.

                   D:  The  reported value is from a dilution.
 ILM05.2                               B-24

-------
                                                         Exhibit  B  —  Section  3
                                                               Form  Instructions
                                                                     Form IIA-IN

3.4.2.2.5.3       M  (Analysis Method) Qualifier.   Specified  entries and  their
                  meanings  are  as  follows:

                  P:     ICP-AES
                  MS:    ICP-MS
                  CV:    Manual  Cold Vapor Atomic Absorption  (AA)
                  AV:    Automated  Cold  Vapor AA
                  AS:    Semi-Automated  Spectrophotometric
                  C:     Manual  Spectrophotometric
                  "  ":   Where no data have been entered
                  NR:    If  the  analyte  is not required to  be analyzed

3.4.2.2.6      A brief physical  description  of  the sample, both before and
               after digestion,  shall be  reported in the fields  for color
                (before and  after),  clarity (before and after), texture,  and
               artifacts.   For  water samples,  report color and clarity.   For
               soil  samples, report color,  texture,  and artifacts.   The
               following descriptive terms are recommended:

               '*•  Color -  red,  blue, yellow,  green, orange, violet, white,
                   colorless, brown, grey,  and black;

               a>  Clarity  - clear,  cloudy,  and opaque; and

               ™=  Texture  - fine  (powdery), medium   (sand),  and  coarse (large
                   crystals or  rocks).

               If artifacts are present,  enter "YES" in the  artifacts  field
               and describe the artifacts in the "Comments"  field.   If
               artifacts are not present,  leave this field blank.  Note  any
               significant  changes that occur during sample  preparation  (i.e.,
               emulsion  formation)  in the "Comments" field.   Enter  any sample-
               specific  comments concerning the analyte results  in  the
               "Comments" field.  Also  document raw  instrument results that
               are less  than minus two  times the CRQL (-2xCRQL)  in  the
               "Comments" field and in  the Sample Delivery Group (SDG)
               Narrative.

3.4.2.2.7      If more than two  additional  analytes  were requested,  submit
               Form  IB-IN as appropriate.

3.4.3    Initial  (ICV) and  Continuing Calibration Verification  (CCV)  [Form
         IIA-IN]

3.4.3.1      Purpose.  This  form is  used to  report analyte  recoveries from
             calibration  verification solutions.

3.4.3.2      Instructions.   Complete the header  information according to  the
             instructions in Exhibit B,  Section 3.3.   Complete  the remainder of
             the  form using  the  following  instructions.

3.4.3.2.1      Enter the ICV Source (12 characters maximum)  and  the CCV  Source
                (12 characters maximum).  Enter sufficient  information  in the
               available 12 spaces to identify the manufacturer  and the
               solution  used.

               Use additional Form(s) IIA-IN if more calibration verification
               sources were used.

3.4.3.2.2      Under "Initial Calibration Verification  True",  enter the  value
                [in micrograms per  Liter (ug/L), to one decimal place]  of the
               concentration of each analyte in the  ICV Solution.

                                      B-25                               ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Form IIA-IN  (Con't)

3.4.3.2.3      Under "Initial Calibration Verification  Found",  enter  the most
               recent  value (in ug/L,  to two  decimal  places),  of the
               concentration of each analyte  measured in the ICV Solution.

3.4.3.2.4      Under "Initial Calibration Verification  %R",  enter the value
                (to the nearest  whole number)  of the percent recovery  computed
               according to the following equation:
                EQ.  1   ICV  Percent Recovery

                                   Found (IC\
                                   True(ICV)
%R =  FOUnd(ICV) x 100
                WHERE,  "True (ICV)" is the true concentration of the analyte in
                the ICV Solution and "Found(ICV)" is the found concentration of
                the analyte in the ICV Solution.

                The values used in EQ. 1 for "True(ICV)" and "Found(ICV)" shall
                be exactly those reported on this form.

 3.4.3.2.5       Under "Continuing Calibration Verification True",  enter the
                value (in ug/L, to one decimal place)  of the concentration of
                each analyte in the CCV Solution.

 3.4.3.2.6       Under "Continuing Calibration Verification Found",  enter the
                value (in ug/L, to two decimal places) of the concentration of
                each analyte measured in the CCV Solution.

                NOTE: The form contains two "Continuing Calibration
                Verification Found" columns.  The column to the left shall
                contain values for the first CCV, and the column to the right
                shall contain values for the second CCV.

 3.4.3.2.7       If more than one Form IIA-IN is required to report multiple
                CCVs, then the column to the left on the second form shall
                contain values for the third CCV, the column to the right shall
                contain values for the fourth CCV, and so on.

 3.4.3.2.8       Under "Continuing Calibration Verification %R", enter the value
                (to the nearest whole number) of the percent recovery computed
                according to the following equation:
                EQ. 2  CCV Percent Recovery

                                   Found(CC
                                    True(CCV)
%R = FQUnd(CCV)  x 100
                WHERE, "True(CCV)" is the true concentration of each analyte,
                and "Found(CCV)" is the found concentration of the analyte in
                the CCV Solution.

                The values used in EQ. 2 for "True(CCV)" and "Found(CCV)" shall
                be exactly those reported on this form.

                NOTE: The form contains two "Continuing Calibration
                Verification %R" columns.  Entries to these columns shall
                follow the sequence detailed above for entries to the
                "Continuing Calibration Verification Found" columns.
 ILM05.2                               B-26

-------
                                                         Exhibit B — Section 3
                                                               Form Instructions
                                                                     Form IIB-IN

3.4.3.2.9      Under "M", enter the method used or "NR", as explained in
               Exhibit B, Section  3.4.2.2.5.3.

3.4.3.2.10     If more than one wavelength/mass is used to analyze an analyte,
               submit additional Form(s)  IIA-IN as appropriate.

3.4.3.2.11     The order of reporting ICVs and CCVs for each analyte shall
               follow the chronological order in which  the standards were run.
               Start with the  first Form  IIA-IN and move from  the left  to the
               right, continuing to the following Form  IIA-INs as appropriate.
               For instance, the first ICV for all analytes shall be reported
               on the first Form IIA-IN.  In a run where three CCVs were
               analyzed, the first CCV shall be reported in the left CCV
               column on the first Form IIA-IN and the  second  CCV shall be
               reported  in the right column of the same form.   The third CCV
               shall be  reported in the left CCV column of the second
               Form IIA-IN.  On the second Form IIA-IN, the ICV column  and  the
               right CCV column shall be  left empty in  this example.  In  the
               previous  example, if a second run for  an analyte was needed,
               the ICV of that run shall  be reported  on a third Form IIA-IN
               and the CCVs follow in the same fashion  as explained before.
               In the case where two wavelengths are  used for  an analyte, all
               ICV and CCV results of one wavelength  from all  runs shall  be
               reported  before proceeding to report the results of the  second
               wavelength used.

3.4.4    CRQL Check Standard [Form IIB-IN]

3.4.4.1     Purpose.  This form is used to report analyte recoveries from
            analyses of  the CRQL Check Standards  (CRIs).

3.4.4.2     Instructions.  Complete the header information according to the
            instructions in Exhibit B, Section 3.3.   Complete  the remainder of
            the  form using the following  instructions.

3.4.4.2.1      Enter the CRQL  Check Standard Source  (12 characters maximum) as
               explained in Exhibit B, Section 3.4.3.2.1.

3.4.4.2.2      Under "CRQL Check Standard True", enter  the value (in ug/L,  to
               one decimal place)  of the  concentration  of each analyte  in the
               CRQL Check Standard that was analyzed  for analytical  samples
               associated with the SDG.

3.4.4.2.3      Under "CRQL Check Standard Initial Found", enter the value (in
               ug/L, to  two decimal places) of the concentration of each
               analyte measured in the CRQL Check Standard analyzed at  the
               beginning of each run.  Concentration  units are ug/L.  If
               applicable, enter the concentration qualifier "J" or  "U" after
               the concentration  (e.g., 1.96J for Lead), as specified in
               Exhibit B, Section  3.4.2.2.5.1.

3.4.4.2.4      Under "CRQL Check Standard Initial %R",  enter the value  (to  the
               nearest whole number) of the percent recovery computed
               according to the following equation:

               EQ. 3  CRQL Check Standard Initial Percent Recovery

                         CRQL Check Standard Initial Found
                              CRQL Check  Standard True
                                                             1UU
                                      B-27                               ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Form III-IN

3.4.4.2.5      Under "CRQL Check Standard Final Found", enter the value  (in
               ug/L, to  two  decimal places)  of  the  concentration of  each
               analyte measured in the CRQL  Check Standard analyzed  at  the end
               of  each run.   If applicable,  enter the  concentration  qualifier
               "J" or "U"  after the concentration  (e.g.,  1.96J  for Lead), as
               specified in  Exhibit B, Section  3.4.2.2.5.1.

3.4.4.2.6      Under "CRQL Check Standard Final %R", enter the value  (to the
               nearest whole number)  of  the  percent recovery computed
               according to  the following equation:

               EQ. 4  CRQL Check Standard Final Percent Recovery

                      _  CRQL Check Standard  Final Found x
                            CRQL Check Standard True


3.4.4.2.7      All percent recovery values reported in EQs. 3 and 4  shall be
               calculated  using the exact true  and  found  values reported on
               this form.

               NOTE:  For every initial  solution reported  there  must  be  a final
               one.  However,  the  opposite  is not true.   If a CRQL Check
               Standard  was  required  to  be  analyzed in the middle of a  run,  it
               shall  be  reported  in the  "Final  Found"  section of this form.

3.4.4.2.8      If  more CRI analyses were required or analyses were performed
               using more  than one wavelength per analyte, submit additional
               Form(s)  IIB-IN as  appropriate.

3.4.4.2.9      The order of  reporting CRIs  for  each analyte shall follow the
               chronological order in which the standards were  run  starting
               with the  first Form IIB-IN and continuing  to the following
                Forms IIB-IN as appropriate.   When multiple wavelengths  are
               used for  one analyte,  all the results of one wavelength  shall
               be  reported before  proceeding to the next  wavelength.

3.4.5    Blanks  [Form  III-IN]

3.4.5.1      Purpose.   This form is used  to  report analyte concentrations found
             in the Initial Calibration  Blank (ICB), Continuing  Calibration
             Blanks (CCB),  and the  Preparation Blank (PB).

3.4.5.2      Instructions.   Complete the  header  information according to the
             instructions in Exhibit B,  Section 3.3.  Complete the remainder  of
             the  form using the following instructions.

3.4.5.2.1      Enter "SOIL"  or "WATER"  as appropriate  as  the matrix  of  the PB.
                No abbreviations or other matrix descriptors may be  used.

3.4.5.2.2      According to the matrix  specified  for the  PB, enter the  PB
                concentration units as "UG/L" for water.or "MG/KG"  for soil.

3.4.5.2.3      Under "Initial Calibration Blank", enter the concentration (in
                ug/L,  to one decimal  place)  of each  analyte  in  the most recent
                ICB, as described in Exhibit B,  Section 3.4.5.2.8, below.

3.4.5.2.4      For each  calibration blank associated with a given method and
                instrument, enter "J"  under  the "C"  qualifier  field  on Form
                III-IN if the absolute value of the  analyte  concentration is
                less than the CRQL for water but greater than  or equal to the
 ILM05.2                              B-28

-------
                                                         Exhibit B — Section 3
                                                              Form Instructions
                                                            Form III-IN  (Con't)

               MDL that was obtained  from direct analysis  (Preparation Method
               "NP1") using that method and  instrument.

               For prepared calibration blanks  (e.g., mercury), the CRQL  for
               water and the MDL for  the preparation method, analysis, and
               instrument shall be used.

               Enter "U" if the absolute value  of the analyte in the blank is
               less than the MDL obtained from  direct analysis or the
               preparation method.

3.4.5.2.5      Under "Continuing Calibration Blank 1", enter the concentration
                (in ug/L, to one decimal place)  of each analyte detected in the
               first required CCB analyzed after the ICB,  as described in
               Exhibit B, Section 3.4.5.2.8, below.  Enter any appropriate
               qualifier, as explained for the  "Initial Calibration Blank", to
               the "C" qualifier column immediately following the "Continuing
               Calibration Blank 1" column.

3.4.5.2.6      If up to three CCBs were analyzed, complete  the columns labeled
               "2" and "3" in accordance with the instructions for the
               "Continuing Calibration Blank 1" column.  If more than three
               CCBs were analyzed, then complete additional Form(s) III-IN as
               appropriate.

3.4.5.2.7      Under "Preparation Blank", enter the concentration in ug/L (to
               three decimal places)  for a water blank, or mg/kg  (to three
               decimal places) for a  soil blank, of each analyte in the PB, as
               described in Exhibit B, Section  3.4.5.2.8,  below.  Evaluate the
               absolute value of the  analyte concentration to determine the
               appropriate concentration qualifier, as explained in Exhibit B,
               Section 3.4.2.2.5.1, and enter the qualifier in the "C" column
               immediately following  the "Preparation Blank" column.

3.4.5.2.8      For all blanks, enter  the concentration (positive or negative)
               for each analyte, if the absolute value of  the concentration is
               greater than or equal  to the  appropriate MDL.  Enter the CRQL
               value for the analyte, if the absolute value of the
               concentration is less  than the appropriate  MDL.

               For example, arsenic has a MDL of 3 ug/L for Preparation Method
               "NP1"  [CRQL for arsenic is 15 ug/L  (water)].  Therefore, a CCB
               instrument reading of  -4.2485 ug/L will be  reported as -4.2J; a
               CCB instrument reading of -2.4356 ug/L will be reported as
               15.OU; a CCB instrument reading  of 4.3586 ug/L will be reported
               as 4.4J; and a CCB instrument reading of 2.1584 ug/L will  be
               reported as 15.OU.

3.4.5.2.9      Under "M", enter the method used, as explained in Exhibit  B,
               Section 3.4.2.2.5.3.

3.4.5.2.10     If more than one wavelength/mass is used to  analyze an analyte,
               submit additional Form(s) III-IN as appropriate.

3.4.5.2.11     The order of reporting ICBs and  CCBs for each analyte shall
               follow the chronological order in which the blanks were run
               starting with the first Form  III-IN and moving from left to
               right and continuing to additional Forms III-IN.  When multiple
               wavelengths are used for the  analysis of one analyte, all  the
               results of one wavelength shall  be reported before proceeding
               to the next wavelength.
                                      B-29                               ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Forms IVA-IN and IVB-IN

3.4.6    ICP-AES and ICP-MS Interference Check Sample  (ICS)  [Forms IVA-IN and
         IVB-IN]

3.4.6.1     Purpose.  These forms are used to report ICS  results  for each ICP-
            AES  or  ICP-MS  instrument used in SDG analyses.

3.4.6.2     Instructions.  Complete the header information according to  the
            instructions in Exhibit B,  Section 3.3.  Complete  the remainder  of
            the  form using the  following instructions.   The  instructions for
            Forms IVA-IN and  IVB-IN are identical except where specified.

3.4.6.2.1      For  "ICP Instrument ID", enter an identifier  that  uniquely
                identifies  a  specific  instrument  within the Contractor
                laboratory.  No  two  ICP  instruments within a  laboratory may
                have the  same  ICP Instrument  ID.

3.4.6.2.2      Enter "ICS  Source"  (12 characters maximum) as explained in
                Exhibit B,  Section 3.4.3.2.1.   For USEPA solutions,  include  in
                the  source  name  a number identifying it (e.g.,  EPA-LV87) .

3.4.6.2.3      Under "True Sol.  A", enter the true concentration  (in ug/L, to
                the  nearest whole number)  of  each analyte present  in Solution
                A.   Enter "0" for each analyte  with no  specified true  value  in
                Solution  A.

3.4.6.2.4      Under "True Sol.  AB",  enter the true concentration (in  ug/L,  to
                the  nearest whole number)  of  each analyte present  in Solution
                AB.   Enter "0" for each analyte  with no specified true value in
                Solution  AB.

3.4.6.2.5       Under "Initial Found Sol. A"  on Form IVA-IN  (ICP-AES),  and
                "Found Sol. A" on Form IVB-IN (ICP-MS), enter the  concentration
                 (positive,  negative,  or zero,  in ug/L,  to the nearest whole
                number)  of each analyte and interferent in the initial analysis
                of Solution A as required in Exhibit D.

3.4.6.2.6       Under "Initial Found Sol. A  %R" on Form IVA-IN  (ICP-AES), and
                "Found Sol. A %R" on Form IVB-IN  (ICP-MS), enter the value (to
                the nearest whole number)  of the percent recovery computed for
                true Solution A greater than zero according to the following
                equation:

                 EQ. 5   Initial  Found  Sol. A  Percent Recovery

                       „_  _ Initial Found Solution A
                                 True Solution A
                                                       inn
                                                     x J.UU
                Leave the field blank if "True Solution A" equals zero.

 3.4.6.2.7      Under "Initial Found Sol. AB" on Form IVA-IN (ICP-AES), and
                "Found Sol. AB" on Form IVB-IN (ICP-MS), enter the
                concentration  (positive, negative, or zero, in ug/L, to one
                decimal place) of each analyte and interferent in the initial
                analysis of Solution AB as required in Exhibit D.
 ILM05.2                               B-30

-------
                                                         Exhibit B  —  Section 3
                                                              Forms  Instructions
                                                Forms IVA-IN  and IVB-IN  (Con't)

3.4.6.2.8      Under "Initial Found Sol. AB %R" on Form IVA-IN  (ICP-AES), and
               "Found Sol. AB %R"  on  Form  IVB-IN  (ICP-MS) ,  enter the value  (to
               the  nearest whole number) of the percent recovery computed for
               True Solution AB greater than  zero according to  the  following
               equation:

               EQ.  6  Initial Found  Sol. AB Percent  Recovery

                         _ Initial Found Solution AB x  QQ
                                True Solution  AB


               Leave the  field blank if "True Solution AB"  equals  zero.

3.4.6.2.9      Under "Final Found  Sol.  A", enter  the concentration  (positive,
               negative,  or  zero,  in ug/L,  to the nearest whole number)  of
               each analyte  and interferent in the final  analysis  of  Solution
               A as required in Exhibit D.  ICP-MS analysis  (Form  IVB-IN) does
               not  require a final analysis.

3.4.6.2.10     Under "Final Found  Sol.  A %R"  enter the value (to the nearest
               whole number) of the  percent recovery computed for  true
               Solution A greater  than zero according  to  the following
               equation:

               EQ.  7  Final  Found  Sol. A Percent  Recovery

                             Final Found Solution  A
                                True Solution A
                                                     .. ._
                                                    Ť .L U U
                Leave the field blank if "True Solution A" equals zero.

3.4.6.2.11      Under "Final  Found Sol.  AB",  enter the  concentration (positive,
                negative,  or  zero, in ug/L,  to one decimal place) of each
                analyte and interferent in the final analysis of Solution AB as
                required in Exhibit D.  ICP-MS analysis (Form IVB-IN)  does not
                require a final analysis.

3.4.6.2.12      For all found values  of Solutions  A and AB,  enter the
                concentration (positive, negative, or zero)  of each analyte and
                interferent at each wavelength used for analysis by ICP.

3.4.6.2.13      Under "Final  Found Sol.  AB %R",  enter the  value (to the  nearest
                whole number) of the percent recovery computed for true
                Solution AB greater than zero according to the following
                equation:

                EQ. 8  Final  Found Sol. AB Percent Recovery

                             Final Found Solution  AB
                                True Solution AB
                                                      1f.ri
                                                      J.UU
                Leave the field empty if "True Solution AB" equals zero.

                All percent recovery values reported shall be calculated using
                the exact true and found values reported on this form.

                NOTE: For ICP-AES (Form IVA-IN), for every initial solution
                reported there must be a final solution reported.  However, the
                opposite is not true.  If an  ICS was required to be analyzed in

                                      B-31                               ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Form VA-IN

                the  middle  of a run,  it  shall be  reported in  the  "Final  Found"
                section  of  this form.

3.4.6.2.14      If more  ICS analyses were required, submit additional Form(s)
                IVA-IN and/or IVB-IN  as  appropriate.

3.4.6.2.15      The  order of  reporting ICSs for each analyte  shall  follow  the
                chronological order in which  the  standards were  run,  starting
                with the first Form IVA-IN  and/or IVB-IN and  continuing  to the
                following Forms IV-IN as appropriate.   When multiple
                wavelengths/masses are used for one analyte,  all  the  results of
                one  wavelength/mass shall be  reported  before  proceeding  to the
                next wavelength/mass.

3.4.7    Matrix Spike Sample Recovery [Form VA-IN]

3.4.7.1      Purpose.  This form is used to report results for the  pre-digest
             spike.

3.4.7.2      Instructions.  Complete  the header information according to the
             instructions in  Exhibit  B,  Section 3.3.   Complete the  remainder of
             the form using the following instructions.

3.4.7.2.1       Indicate the  appropriate matrix,  level,  and concentration units
                 (ug/L for water and mg/kg dry weight  for soil)  as explained in
                Exhibit  B,  Sections  2.5.2.1.1 and 3.3.8.

3.4.7.2.2       For  "% Solids for Sample",  enter  the percent  solids (see
                Exhibit  B,  Section 3.4.2.2.2)  for the  original  sample of EPA
                sample  number reported on the form.  Note that  this number must
                equal the one reported on Form IA-IN  for that sample.

3.4.7.2.3       In the "EPA Sample No."  box,  enter an  EPA sample  number  (7
                places  maximum) of the sample from which the  spike results on
                this form were obtained.  The number  shall be centered in the
                box.

3.4.7.2.4       Under "Control Limit  %R",  enter "75-125" if the  sample  result
                is less  than or equal to four times the spike added value.  If
                the sample  result is  greater  than four times  the Spike Added
                 (SA) value, leave this field empty.

3.4.7.2.5       Under "Spiked Sample  Result (SSR).", enter the measured value
                 (to four decimal places), in  appropriate units,  for each
                relevant analyte in the  matrix spike  sample.   Enter any
                appropriate concentration qualifier,  as explained in Exhibit B,
                Section 3.4.2.2.5.1,  to  the "C" qualifier column immediately
                following the "Spiked Sample  Result (SSR)" column.

3.4.7.2.6       Under "Sample Result  (SR)", enter the  measured  value  (to four
                decimal places) for each required analyte in the sample
                 (reported in "EPA Sample No." box) on which the matrix spike
                was performed.  Enter any appropriate concentration qualifier,
                as explained in Exhibit B,  Section 3.4.2.2.5.1,  to the "C"
                qualifier column immediately following the "Sample Result  (SR)"
                column.

3.4.7.2.7       Under "Spike Added (SA)", enter the value (to two decimal
                places)  for  the concentration of each analyte added to the
                sample.   The same concentration units shall be used for "SSR",
                 "SR", and "SA".  If the "Spike Added" concentration is
                 specified in the contract,  the value added and reported shall

 ILM05.2                              B-32

-------
                                                         Exhibit B — Section 3
                                                              Form Instructions
                                                                      Form VB-IN

               be the specific concentration in appropriate units, corrected
               for spiked sample weight and percent solids  (soils) or spiked
               sample volume  (waters).

3.4.7.2.8      Under "%R", enter the value (to the nearest whole number) of
               the percent recovery  for all spiked analytes computed according
               to the following equation:

               EQ. 9  Spike Percent  Recovery

                               %R =  SSR - SR x 100
                                       SA


               Percent recovery shall  be reported, whether it is negative,
               positive or zero.

               The values for  "SSR",  "SR", and "SA" must be exactly those
               reported on this form.  A value of zero  shall be used  in
               calculations for "SSR"  or "SR" if the analyte value is less
               than the MDL.

3.4.7.2.9      Under "Q", enter "N"  if the Spike Recovery (%R) is out of the
               control limits  (75-125) and the Sample Result  (SR) is  less  than
               or equal to four times  the SA.

3.4.7.2.10     Under "M", enter the  method used (as explained in Exhibit B,
               Section 3.4.2.2.5.3)  or enter "NR" if the analyte is not
               required in the spike.

3.4.7.2.11     If different samples  were used for spike sample analysis of
               different analytes, additional Form(s) VA-IN shall be  submitted
               for each sample as appropriate.

3.4.8    Post-Digestion Spike Sample Recovery [Form VB-IN]

3.4.8.1     Purpose.  This form is used to report results for the post-digest
            spike recovery which is  based upon the addition of a known
            quantity of analyte to an  aliquot of the digested  sample.

3.4.8.2     Instructions.  Complete  the header information according to the
            instructions in Exhibit  B, Section 3.3.  Complete the remainder of
            the form using the following instructions.

3.4.8.2.1      In the "EPA Sample No." box, enter an EPA sample number  (seven
               characters maximum) of  the sample from which the spike results
               on this form were obtained.  The number  shall be centered  in
               the box.

3.4.8.2.2      The "Control Limit %R"  and "Q" fields shall be left blank until
               limits are established  by USEPA.  At that time, the Contractor
               will be informed how  to complete these fields.

3.4.8.2.3      Under "Spiked Sample  Result (SSR)",  enter the measured value
                (in ug/L, to two decimal places) for each analyte in the post-
               digest spike sample.  Enter any appropriate concentration
               qualifier, as explained in Exhibit B, Section 3.4.2.2.5.1,  to
               the "C" qualifier column immediately following the "Spiked
               Sample Result  (SSR)"  column.

3.4.8.2.4      Under "Sample Result  (SR)", enter the measured value (in ug/L,
               to two decimal places)  for the concentration of each analyte in

                                      B-33                              ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Form VI-IN

                the  sample  (reported  in  "EPA Sample No." box) on  which  the
                spike was performed.   Enter  any  appropriate  concentration
                qualifier,  as  explained  in Exhibit B,  Section 3.4.2.2.5.1,  to
                the  "C"  qualifier  column immediately  following  the  "Sample
                Result  (SR)" column.

3.4.8.2.5       Under "Spike Added (SA)",  enter  the value  (in ug/L, to  one
                decimal  place)  for each  analyte  added to the sample.   If the SA
                concentration  is specified in the contract,  the value added and
                reported shall be  that specific  concentration in  appropriate
                units.

3.4.8.2.6       Under "%R", enter  the value  (to  the nearest  whole number) of
                the  percent recovery  for all spiked analytes computed according
                to EQ.  9 in Exhibit B, Section 3.4.7.2.8.   Percent  recovery
                shall be reported, whether it is negative,  positive,  or zero.
                The  values  for "SSR", "SR",  and  "SA"  must  be exactly  those
                reported on this form.  A value  of  zero shall be  substituted
                for  "SSR"  or "SR"  if  the analyte value is  less  than the MDL.

3.4.8.2.7       Under "M",  enter the  method  used as explained in  Exhibit B,
                Section 3.4.2.2.5.3,  or  enter "NR"  if the  spike was not
                required.

3.4.8.2.8       If different samples  were used for  spike sample analysis of
                different analytes, additional Form(s) VB-IN shall  be
                submitted.

3.4.9     Duplicates [Form  VI-IN]

3.4.9.1      Purpose.   The  duplicates form is used  to report results  of
             duplicate  analyses.  Duplicate  analyses  are  required for percent
             solids  values  and all analyte results.

3.4.9.2      Complete the header information according to  the instructions  in
             Exhibit B,  Section 3.3.   Complete the  remainder of the form using
             the following instructions.

3.4.9.2.1       Indicate the appropriate matrix, level, and concentration units
                (ug/L for water and mg/kg dry weight for soil)  as explained in
                Exhibit B,  Sections 2.5.2.1.1 and 3.3.8.

3.4.9.2.2       For  "%  Solids  for Sample", enter the  percent solids (as
                explained in Exhibit B,  Section 3.4.2.2.2) for the original
                sample of the EPA sample number reported on the form.  Note
                that this number must equal the one reported on Form IA-IN for
                that sample.

3.4.9.2.3      For  "%  Solids  for Duplicate", enter the percent solids  (as
                explained in Exhibit B,  Section 3.4.2.2.2) for the duplicate
                sample of the EPA sample number reported on the form.

3.4.9.2.4      In the  "EPA Sample No."  box, enter  EPA sample  number  (seven
                characters maximum)  of  the  sample from which the duplicate
                sample results  on this  form were obtained.  The number  shall be
                centered in the box.

3.4.9.2.5      Under "Control Limit", enter the CRQL  (in appropriate units,
                ug/L for water  or mg/kg dry weight basis corrected for  the
                original sample weight  and  percent solids) for the analyte if
                either the sample or  duplicate value was less than 5 times the
                CRQL.  If the sample  and duplicate values were greater  than or

 ILM05.2                              B-34

-------
                                                         Exhibit B — Section 3
                                                              Form Instructions
                                                                     Form VII-IN

               equal to 5 times the CRQL, or if the sample and duplicate
               values were less than the CRQL, leave the field empty.

3.4.9.2.6      Under "Sample  (S)", enter the original measured value  (to four
               decimal places) for the concentration of each analyte  in  the
               sample  (reported in "EPA Sample No." box) on which a  duplicate
               analysis was performed.  Concentration units are those
               specified on the form.  Enter any appropriate concentration
              . qualifier, as  explained in Exhibit B, Section 3.4.2.2.5.1, to
               the "C" qualifier  column immediately following the "Sample  (S)"
               column.

3.4.9.2.7      Under "Duplicate (D)", enter the measured value (to four
               decimal places) for each analyte in the duplicate sample.
               Concentration  units are those specified on the form.   Enter any
               appropriate concentration qualifier, as explained in  Exhibit B,
               Section 3.4.2.2.5.1, to the "C" qualifier column immediately
               following the  "Duplicate  (D)" column.

3.4.9.2.8      For solid samples,  the concentration of the original sample
               shall be computed  using the weight and percent solids  of  the
               original sample.   The concentration of the duplicate  sample
               shall be computed  using the weight of the duplicate sample, but
               the percent solids of the original sample.

3.4.9.2.9      Under "RPD", enter the absolute value (to the nearest whole
               number) of the Relative Percent Difference  (RPD) for  all
               analytes detected  above the MDL in either the sample  or the
               duplicate, computed according to the following equation:

               EQ. 10  Duplicate  Sample Relative Percent Difference

                               RPD =   I S "  D I   x 100
                                     (S + D)/2


               The values for "S" and "D" shall be exactly those reported on
               this form.  A  value of zero shall be substituted for  "S"  or "D"
               if the analyte concentration is less than the MDL in  either
               one.  If the analyte concentration is less than the MDL in both
               "S" and "D", leave the "RPD" field empty.

3.4.9.2.10     Under "Q", enter "*" if the duplicate analysis for the analyte
               is out of control.  If both sample and duplicate values are
               greater than or equal to 5 times the CRQL, then the RPD must be
               less than or equal to 20% to be in control.  If either the
               sample or duplicate value is less than 5 times the CRQL,  then
               the absolute difference between the sample and duplicate values
               shall be less  than the CRQL to be in control.

3.4.9.2.11     If both values are below the CRQL,  then no control limit is
               applicable.

3.4.9.2.12     Under "M", enter method used as explained in Exhibit B, Section
               3.4.2.2.5.3.

3.4.10   Laboratory Control Sample [Form VII-IN]

3.4.10.1    Purpose.  This form is used to report results for the solid and
            aqueous LCSs.
                                      B-35                               ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Form VII-IN  (Con't)

3.4.10.2     Instructions.  Complete the header information according to the
             instructions  in  Exhibit B, Section 3.3.  Complete  the  remainder  of
             the  form using the  following  instructions.

3.4.10.2.1     For  the  Solid LCS  Source  (12  characters maximum), enter the
               appropriate EPA  sample number if  EPA provided standard  was
               used.   Substitute  an appropriate  number provided by EPA for LCS
               solutions  prepared in  the  future.   If other  sources were  used,
               identify the  source.   For  the aqueous LCS Source, enter the
               source  name (12  characters maximum)  as explained in Exhibit B,
               Section 3.4.3.2.1.

3.4.10.2.2     Under "Aqueous True",  enter the value (in ug/L,  to  one  decimal
               place)  of  the concentration of each analyte  in the  Aqueous LCS
               Standard Source.

3.4.10.2.3     Under "Aqueous Found",  enter  the  measured concentration (in
               ug/L,  to two  decimal places)  of each analyte found  in the
               Aqueous LCS solution.

3.4.10.2.4     Under "Aqueous %R", enter  the value of the percent  recovery  (to
                the nearest whole number)  computed according to the following
                equation:

                EQ. 11   Aqueous  LCS Percent Recovery

                                Aqueous LCS Found
                           %R
                                Aqueous LCS True
 3.4.10.2.5     Under "Solid True",  enter the value (in mg/kg,  to one decimal
                place) of the concentration of each analyte in the solid LCS
                Source.

 3.4.10.2.6     Under "Solid Found", enter the measured value (in mg/kg,  to one
                decimal place) of each analyte found in the solid LCS solution.

 3.4.10.2.7     Under "C",  enter "J" or "U" or leave empty,  to describe the
                found value of the  solid LCS as explained in Exhibit B, Section
                3.4.2.2.5.1.

 3.4.10.2.8     Under "Limits",  enter the lower limit (in mg/kg,  to one decimal
                place) in the left  column, and the upper limit (in mg/kg, to
                one decimal place)  in the right column, for each analyte in the
                solid LCS solution.

 3.4.10.2.9     Under "Solid %R", enter the value of the percent recovery (to
                the nearest whole number) computed according to the following
                equation:

                EQ. 12  Solid LCS Percent Recovery

                                 Solid LCS Found
                            %R
                                  Solid LCS True
 3.4.10.2.10    The values for true and found aqueous and solid LCSs used in
                EQs. 11 and  12 shall be exactly those reported on this form.
                If the analyte concentration is less than the MDL, a value of
                zero shall be substituted for the aqueous and solid LCS found.
 ILM05.2                               B-36

-------
                                                         Exhibit B -- Section 3
                                                              Form Instructions
                                                                   Form VIII-IN

3.4.10.2.11    Submit additional Form(s) VII-IN as appropriate if more  than
               one aqueous LCS or solid LCS was required.

3.4.11   ICP-AES and ICP-MS Serial Dilutions  [Form VIII-IN]

3.4.11.1    Purpose.  This form is used to report results for ICP-AES and ICP-
            MS serial dilutions.

3.4.11.2    Instructions.  Complete the header information according to the
            instructions in Exhibit B,  Section 3.3.  Complete the remainder of
            the form using the following instructions.

3.4.11.2.1     In the "EPA Sample No." box, enter EPA sample number  (7  places
               maximum) of the sample  for  which serial  dilution analysis
               results on this form were obtained.   The number shall be
               centered in the box.

3.4.11.2.2     Under "Initial Sample Result (I)", enter the instrument
               measured value  (in ug/L to  two decimal places) for each  ICP
               analyte.  This value shall  not be corrected for any dilution.
               Enter any appropriate concentration qualifier, as explained in
               Exhibit B, Section 3.4.2.2.5.1, to the "C" qualifier  column
               immediately following the "Initial Sample Result  (I)" column.

               NOTE: The initial sample concentration for an analyte does not
               have to equal the value for that analyte reported on  Form IA-IN
               for that sample.  It is the value of  the analyte's instrument
               measured value  (uncorrected for dilution) that is within the
               linear range of the instrument.

3.4.11.2.3     Under "Serial Dilution  Result  (S)", enter the instrument
               measured value corrected for a five-fold dilution  (in ug/L to
               two decimal places) for each ICP analyte in the diluted  sample.
               Enter any appropriate concentration qualifier, as explained in
               Exhibit B, Section 3.4.2.2.5.1, to the "C" qualifier  column
               immediately following the "Serial Dilution Result  (S)" column.

               NOTE: The "Serial Dilution  Result  (S)" is obtained by
               multiplying by five the instrument measured value  (in ug/L) of
               the serially diluted sample.   The "C" qualifier for the  serial
               dilution shall be established  based on the serial dilution
               result before correcting it for the five-fold dilution
               regardless of the value reported on Form VIII-IN.

               For example, if the instrument readout value for the  "Initial
               Sample Result  (I)" for  silver  in a two-fold diluted sample
               MAX123 is 1164.36 ug/L,  and the instrument readout value for
               the "Serial Dilution Result (S)" for  silver in a ten-fold
               diluted sample MAX123  (MAX123L) is 241.67 ug/L, then  the
               concentration reported  for  silver in  the "Initial Sample Result
                (I)" column will be 1164.36 ug/L  (not 2  times the instrument
               readout value which equals  2328.72 ug/L), and the concentration
               reported for silver in  the  "Serial Dilution Result  (S)"  column
               will be five times the  instrument readout value which equals
               1208.35 ug/L  (not 10 times  the instrument readout value  which
               equals 2416.70 ug/L).
                                      B-37                               ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Form IX-IN

3.4.11.2.4     Under "% Difference", enter the absolute value  (to  the nearest
               whole number)  of the percent  difference in concentration of
               required analytes,  between the original sample  and  the diluted
               sample  (adjusted for dilution)  according to the following
               formula:

               EQ.  13  Serial Dilution Percent Difference
                                          I T  _ q I
                           % Difference =  *-ą	fj
                The values for "I" and "S" used to calculate percent difference
                in EQ.  13 shall be exactly those reported on this form.   A
                value of zero shall be substituted for "S" if the analyte
                concentration is less than the MDL.  If the analyte
                concentration in  (I) is less than the MDL concentration,  leave
                the "% Difference" field empty.

 3.4.11.2.5      Under "Q",  enter "E" if the percent difference is greater than
                10% and the original sample concentration (reported on Form IA-
                IN) is greater than 50 times the MDL reported on Form IX-IN.

 3.4.11.2.6      Under "M",  enter the method of analysis for each analyte  as
                explained in Exhibit B, Section 3.4.2.2.5.3.

 3.4.12    Method Detection Limits (Annually)  [Form IX-IN]

 3.4.12.1     Purpose.  This form documents the Method Detection Limits (MDLs)
             for each preparation method and instrument that the Contractor
             used to obtain data for the SDG.  Only the methods, instruments,
             and wavelengths used to generate data for the SDG shall be
             included.  The Contractor shall also report MDLs, obtained from
             direct analysis, for each instrument used to obtain data for the
             SDG.  The MDLs obtained from direct analysis shall be used in the
             qualification of data associated with samples and instrument QC
             standards that are hot taken through a preparation procedure.
             Although the MDLs are determined annually, a copy of the annual
             MDLs shall be included with each Sample Data Package on Forms IX-
             IN.

 3.4.12.2     Instructions.  Complete the header information according to  the
             instructions in Exhibit B, Section 3.3.  Complete the remainder of
             the form using the following instructions.

 3.4.12.2.1      Enter the Analysis Method qualifier as specified in Exhibit B,
                Section 3.4.2.2.5.3, in the "Instrument Type" field.

 3.4.12.2.2      Enter the Instrument ID in the "Instrument ID" field (12
                characters maximum).   These instrument IDs are used to uniquely
                identify each instrument that  the  laboratory used to perform
                the  analysis.

 3.4.12.2.3      Enter the date  (formatted MM/DD/YYYY) on which the MDL analysis
                was  performed in  the "Date" field.
 ILM05.2                               B-38

-------
                                                         Exhibit B — Section 3
                                                              Form Instructions
                                                             Form IX-IN  (Con't)

3.4.12.2.4     For "Preparation Method", enter the method of preparation
                (three characters maximum) for which the MDLs listed on Form
                IX-IN were established.  Use appropriate sample preparation
                codes as specified below:

                HW1:  Hotplate/Block  digestion  for ICP-AES  analysis of  water
                     samples.
                HW2:  Hotplate/Block  digestion  for ICP-MS analysis of water
                     samples.
                MW1:  Microwave  digestion for ICP-AES analysis of water
                     samples.
                MW2:  Microwave  digestion for ICP-AES analysis of water
                     samples.
                HS1:  Hotplate/Block  digestion  for ICP-AES  analysis of  soil
                     samples.
                HS2:  Hotplate/Block  digestion  for ICP-AES  analysis of  soil
                     samples.
                MSI:  Microwave  digestion for ICP-AES analysis of soil  samples.
                CW1:  Preparation for the Manual  Cold Vapor AA analysis of
                     water samples.
                CS1:  Preparation for the Manual  Cold Vapor AA analysis of soil
                     samples.
                CW2:  Preparation for the Automated Cold Vapor AA analysis of
                     water samples.
                DW1:  Distillation for the manual and semi-automated  spectro-
                     photometric analysis of water samples.
                DW2:  Midi-distillation for the semi-automated spectro-
                     photometric analysis of water samples.
                DS1:  Distillation for the manual and semi-automated  spectro-
                     photometric analysis of  soil samples.
                DS2:  Midi-distillation for the semi-automated spectro-
                     photometric analysis of  soil samples.
                NP1:  No preparation.

3.4.12.2.5      Enter the concentration  units  (UG/L for water or MG/KG wet
                weight for soil)  for the results  reported  on Form IX-IN in the
                "Concentration  Units" field.   Enter "UG/L"  for MDL results
                obtained from direct analysis  (Preparation Method "NP1").

3.4.12.2.6      Under "Wavelength/Mass", enter the wavelength in nanometers
                (nm)  to two  decimal  places or  the mass  in  atomic mass  units
                (amu) to two decimal places for  each analyte for which an MDL
                has been established and is listed in the  MDL column.  If more
                than  one wavelength  or mass is used for an -analyte, use
                additional Form(s) IX-IN as appropriate to  report the  MDL.

3.4.12.2.7      Contract Required Quantitation Limits (in ug/L for water  and
                mg/kg for  soil)  as established in Exhibit  C, shall be  reported
                in  the column headed "CRQL".   The CRQL  shall be reported  in
                ug/L  on Form(s)  IX-IN associated with Preparation Method  "NP1".

3.4.12.2.8      Under "MDL", enter the MDL  (in ug/L for water and direct
                analysis,  or mg/kg for soil, to  two significant figures for
                values less  than 10,  and three significant  figures for values
                greater than or equal to 10) as  determined by the Contractor
                for each analyte analyzed by the  instrument for which  the ID is
                listed on  this  form.   When calculating  MDL values, always round
                up  to the  appropriate significant figure  (e.g., 14.81  rounds to
                14.9  and 146.6  rounds to 147).   This deviation from the
                rounding rule is necessary to  prevent the  reporting of detected
                values for results that  fall in  the noise  region of the
                calibration  curve.

                                      B-39                               ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Form XA-IN

               NOTE:  Zeros  used to set  the  decimal  point  in  a number less  than
               one  are  not  significant  but  all  trailing zeros are significant.

               For  example,  a calculated MDL  value  of  0.074  ug/L will be
               reported as  0.074 and a  calculated MDL  value  of 0.1 or 0.08
               will be  reported as 0.10 and 0.080,  respectively.

3.4.12.2.9     Use  additional  Form(s) IX-IN if more preparation  methods,
               instruments  and wavelengths  or masses are  used.  Note that  the
               date on  this form shall  not  exceed the  analysis dates in the
               Sample Data  Package or precede them  by  more than twelve months.

3.4.12.2.10    Use  the  "Comments"  section to  indicate  alternative wavelengths
               and  the  conditions under which they  are used.

3.4.13    ICP-AES  Interelement Correction Factors  (Quarterly)  [Form XA-IN]

3.4.13.1     Purpose.  This  form documents for each  ICP-AES instrument the
             interelement correction factors applied by the Contractor to
             obtain  data for the SDG. Although  the  correction factors are
             determined  quarterly,  a copy of the results of the quarterly
             interelement correction factors shall be  included with each Sample
             Data  Package on Form XA-IN  and  Form XB-IN  as  appropriate.

3.4.13.2     Instructions.   Complete the header  information according to
             instructions in Exhibit B,  Section  3.3.  Complete the remainder  of
             the  form  using  the following instructions.

3.4.13.2.1     Enter  the ICP-AES Instrument ID  (12  characters maximum), which
                is a unique  number designated  by the Contractor to identify
                each ICP-AES instrument used to  produce data in the Sample  Data
                Package.  If more than one ICP-AES instrument is used, submit
                additional Form(s) XA-IN as  appropriate.

3.4.13.2.2     Report the date  (formatted as  MM/DD/YYYY)  on which these
                correction factors were determined for use.  This date shall
                not exceed the ICP-AES analysis  dates in the Sample Data
                Package  or precede them by more  than three calendar months.

3.4.13.2.3     Under  "Wavelength", list the wavelength in nm  (to two decimal
                places)  used for each ICP-AES  analyte.   If more than one
                wavelength is used, submit additional Form(s) XA-IN or Form(s)
                XB-IN as appropriate.

3.4.13.2.4     Under  "Al",  "Ca", "Fe",  and  "Mg",  enter the correction factor
                (negative, positive or zero, to  seven decimal places, 10
                characters maximum) for each ICP-AES analyte.  Correction
                factors for one  other analyte  shall be reported using the empty
                column and listing the analyte's chemical  symbol  in the blank
                two-space header field provided for that column.

3.4.13.2.5      If corrections are not applied for an analyte, a zero shall be
                entered for that analyte  to indicate that  the  corrections were
                determined to be zero.  Correction  factors for more than one
                additional analyte shall be reported using Form XB-IN.

                NOTE:  Correction factors  for Al, Ca, Fe, and Mg are all
                required and are to be listed first (as they appear on Form XA-
                IN)  .
 ILM05.2                               B-40

-------
                                                           Exhibit B—Section  3
                                                              Form Instructions
                                                  Form XB-IN, XI-IN,  and XII-IN

3.4.14   ICP-AES Interelement Correction Factors  (Quarterly)  [Form XB-IN]

3.4.14.1    Purpose.  This form is used if correction factors for analytes
            other than Al, Ca, Fe, Mg, and one more analyte  of the
            Contractor's choice were applied to the analytes analyzed by ICP-
            AES.

3.4.14.2    Instructions.  Complete this form following the  instructions for
            Form XA-IN  (see Exhibit B, Section 3.4.13) by listing the chemical
            symbol  for additional analytes in the heading of the empty columns
            in the  two-space  fields provided.

3.4.14.2.1     Columns of correction factors for additional  analytes shall be
               entered left to right starting on  Form XA-IN  and proceeding to
               Form XB-IN, according to the alphabetical order of their
               chemical  symbols.

3.4.15   ICP-AES and ICP-MS Linear Ranges  (Quarterly) [Form  XI-IN]

3.4.15.1    Purpose.  This form documents the quarterly linear range analysis
            for each ICP instrument that the Contractor used to obtain data
            for the SDG.

3.4.15.2    Instructions.  Complete the header information according to the
            instructions in Exhibit B, Section 3.3.  Complete the remainder of
            the form using the following instructions.

3.4.15.2.1     Enter the ICP  Instrument ID  (12 characters maximum), which is a
               unique number  designated by  the Contractor to identify each ICP
               instrument used to produce data for  the SDG.   If more than one
   \            ICP  instrument is used, submit additional Form(s) XI-IN as
               appropriate.

3.4.15.2.2     Report the date (formatted as MM/DD/YYYY) on  which these linear
   ,<           ranges were analyzed.   This  date shall not exceed the dates of
               analysis  by ICP in the  Sample Data Package and shall not
               precede  the analysis dates by more than three calendar months.

3.4.15.2.3     Under "Integ.  Time (Sec.)", enter  the integration time (in
               seconds  to two decimal  places) used  for each  measurement  taken
               from the  ICP instrument.

3.4.15.2.4     Under "Concentration",  enter the concentration (in ug/L)  that
               is the upper limit of the ICP instrument linear range as
               determined in  Exhibit D.  Any measurement above it is out of
               the  linear range, and thus,  is an  estimated value and shall be
               diluted  into the linear range.

3.4.15.2.5     Under "M", enter the method of analysis for each analyte as
               explained in Exhibit B, Section 3.4.2.2.5.3.

3.4.15.2.6     If more instruments or  analyte wavelengths/masses are used,
               submit additional Form(s) XI-IN as appropriate.

3.4.16   Preparation Log [Form XII-IN]

3.4.16.1    Purpose.  This form is used to report the preparation run log.

3.4.16.1.1     All  field samples and all Quality  Control (QC) preparations
                (including duplicates,  matrix spikes, LCSs,  PBs, and re-
               preparations)  associated with the  SDG shall be reported on
               Form XII-IN.

                                      B-41                               ILM05.2

-------
Exhibit B -- Section 3
Form Instructions
Form XIII-IN

3.4.16.1.2     Submit one Form XII-IN per batch, per method, if no more than
               thirty-two preparations,  including  QC preparations, were
               performed.   If more  than  32  preparations  per batch, per method,
               were  performed, then submit  additional  copies of Form XII-IN as
               appropriate.  Submit a  separate  Form XII-IN for each  batch.

3.4.16.1.3     The order in which the Preparation  Logs are submitted is very
               important.   Form  XII-IN shall  be organized by method,  by batch.
               Later batches within a method  shall follow earlier ones.   Each
               batch shall  start on a  separate  Form XII-IN.

3.4.16.2    Instructions.  Complete the  header  information according to the
            instructions in Exhibit B, Section  3.3.  Complete the remainder of
            the  form using  the following instructions.

3.4.16.2.1     For "Preparation  Method",  enter  the method of preparation
                (three characters maximum)  for which the  preparations listed on
               Form  XII-IN  were  made,  as specified in  Exhibit B,  Section
               3.4.12.2.4.

3.4.16.2.2     Under "EPA Sample No.", enter  EPA sample  number of each sample
               in the SDG,  and of all  other preparations such as  duplicates,
               matrix spikes,  LCSs, PBs, and  re-preparations (all formatted
               according to Exhibit B,  Table  2).  All  EPA sample  numbers  shall
               be listed in ascending  alphanumeric order, continuing to  the
               next  Form XII-IN  if applicable.

3.4.16.2.3     Under "Preparation Date",  enter  the date  (formatted MM/DD/YYYY)
               on which each sample was  prepared for  analysis by  the method
               indicated in the  header section  of  the  form.

               NOTE: The date  never changes on  a single  Form XII-IN because
               the form shall  be submitted per  batch.

3.4.16.2.4     Under "Weight", enter the wet  weight (in  grams, to two decimal
               places)  of each soil sample prepared for  analysis  by the  method
                indicated in the  header section  of  the  form.  If  the sample
               matrix is water,  then leave the  field empty.

3.4.16.2.5     Under "Volume", enter the final  volume  (in mL, to  the nearest
                whole number)  of  the preparation for each sample  prepared for
                analysis by the method indicated in the header section of the
                form.  This field shall have a value for each sample listed.

3.4.17    Analysis Run Log  [Form  XIII-IN]

3.4.17.1     Purpose. This  form  is used to report  the  sample analysis run log.

3.4.17.1.1     A run is defined  as the totality of analyses performed by  an
                instrument throughout the sequence initiated by,  and including,
                the first SOW-required calibration standard or tune standard,
                and terminated by,  and including, the CCV and CCB following the
                last SOW-required analytical sample.

3.4.17.1.2     All field samples and all QC analyses  (including  tunes,
                calibration standards,  ICVs, CCVs,  ICBs,  CCBs,  CRIs,  ICSs,
                LRSs, LCSs,  PBs,  duplicates, serial dilutions,  matrix spikes,
                and post-digestion/distillation spikes) associated with the SDG
                shall be reported on Form XIII-IN.   The run shall be continuous
                and inclusive of all analyses performed on  the particular
                instrument during the run.
 ILM05.2                               B-42

-------
                                                         Exhibit B — Section 3
                                                              Form Instructions
                                                           Form XIII-IN  (Con't)

3.4.17.1.3     Submit one Form XIII-IN per run if no more than thirty-two  (32)
               analyses, including instrument  calibration, were analyzed in
               the run.  If more than thirty-two analyses were performed in
               the run, submit additional Form(s) XIII-IN as appropriate.

3.4.17.1.4     The order in which the Analysis Run Logs are submitted is very
               important.  Form XIII-IN  shall  be organized by method, and  by
               run.  Later runs within a method shall  follow earlier ones.
               Each analytical run shall start on a  separate Form XIII-IN.
               Therefore, instrument calibration or  tune shall be the first
               entry on the form for each new  run.   In addition, the run is
               considered to have ended  if it  is interrupted for any reason,
               including termination for failing QC  parameters.

3.4.17.2    Instructions.  Complete the  header information according to  the
            instructions in Exhibit B, Section 3.3.  Complete the remainder of
            the  form using the following instructions.

3.4.17.2.1     For "Instrument ID", enter the  Instrument ID  (12 characters
               maximum) which shall be an identifier designated by the
               Contractor to uniquely identify each  instrument used to produce
               data which are required to be reported  in the SDG deliverable.
               If more than one instrument is  used,  submit additional Form(s)
               XIII-IN as appropriate.   The  Instrument ID shall exactly  match
               that reported on Forms IVA, IVB, IX,  XA, XB, XI, XIV, and XV.

3.4.17.2.2     For "Analysis Method", enter the method code  (two characters
               maximum) according to the specifications in Exhibit B, Section
               3.4.2.2.5.3.

3.4.17.2.3     For "Start Date", enter the date (formatted MM/DD/YYYY) on
               which the analysis run was started.

3.4.17.2.4     For "End Date", enter the date  (formatted MM/DD/YYYY) on  which
               the analysis run was ended.

3.4.17.2.5     Under "EPA Sample No.", enter EPA sample number of each
               analysis, including all QC operations applicable to the SDG
                (formatted according to Exhibit B, Table 2).  All EPA sample
               numbers shall be listed in increasing chronological  (date and
               time) order of analysis,  continuing to  the next Form XIII-IN
               for the instrument run, if applicable.   The analysis date and
               time of other analyses not associated with the SDG, but
               analyzed by the instrument in the reported analytical run,
               shall be reported.  Those analyses shall be identified with EPA
               sample number of "ZZZZZZ".

3.4.17.2.6     Under "D/F", enter the dilution factor  (to two significant
               figures) by which the final digestate or distillate needed to
               be diluted for each analysis  to be performed.  The dilution
               factor does not include the dilution,  inherent in the
               preparation as specified  by the preparation procedures in
               Exhibit D.

3.4.17.2.7     The dilution factor is required for all  entries on Form XIII-
               IN.

               NOTE: For a particular sample a dilution factor of "1.0"  shall
               be entered if the digestate or  distillate was analyzed without
               adding any further volume of  dilutant or any other solutions to
               the "Volume" or an aliquot of the "Volume" listed on Form XII-
               IN for that sample.

                                      B-43                               ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Form XIV-IN

3.4.17.2.8     For USEPA supplied solutions such as ICVs, ICSs, and LCSs, a
               dilution  factor  shall be entered if the  supplied solution had
               to be diluted to a dilution different  from that specified by
               the instructions provided with the solution.   The dilution
               factor  reported  in such a case shall be  that  which  would make
               the reported true values on the appropriate form for the
               solution  equal those that were supplied  with  the solution by
               USEPA.  For instance, ICV-2(0887) has  a  true  value  of  104.0
               ug/L at a 20-fold dilution.  If the solution  is -prepared at a
               40-fold dilution, a dilution factor of "2.0"  shall  be  entered
               on Form XIII-IN  and the uncorrected instrument reading is
               compared  to a true value of 52 ug/L.   In this example.  Form
               IIA-IN  will have a true value of 104.0 regardless of the
               dilution  used.  The found value for the  ICV shall be corrected
               for the dilution listed on Form XIII-IN  using the following
               formula:

               EQ. 14  ICV/CCV  Correction for Dilution

               Found value on Form II = Instrument readout (ug/L) x D/F

3.4.17.2.9     Under "Time",  enter the time  (in military format -  HHMM) at
               which  each  analysis was performed.

3.4.17.2.10    Under "Analytes", enter WX" in the column of  the designated
               analyte to  indicate that the analyte  value was used from  the
               reported  analysis to report data in the  SDG.   Leave the column
               empty  for each analyte  if  the analysis was not used to report
               the particular analyte.

3.4.17.2.11    Entering  "X" appropriately is very important.  The  "X" is used
               to link the samples with their related QC.   It also links  the
               dilution  factor with the appropriate  result  reported on
               Inorganic Forms  I-VIII.  For  each  analyte result  reported on
               any  of the  Forms I-VIII, there shall  be  one,  and  only  one,
               properly  identified entry  on  Form  XIII-IN for which an "X"  is
               entered in  the column  for  that analyte.

3.4.17.2.12    If, on  Form XIII-IN, an "X" is entered in the column for an
               analyte for a field sample associated with  a  dilution  factor
               greater than  1.0, flag  the data  for  that analyte  with  a "D"  on
                the  appropriate Form  IA-IN or  Form IB-IN.

3.4.18    ICP-MS  Tune  [Form XIV-IN]

3.4.18.1     Purpose.   This form is used to report the tuning results  for each
             ICP-MS  instrument used in  SDG analyses.

3.4.18.2     Instructions.   Complete the header information according  to  the
             instructions in Exhibit B,  Section  3.3.   Complete the  remainder  of
             the  form using the following  instructions.

3.4.18.2.1     For  "ICP-MS Instrument  ID", enter  an  identifier that uniquely
                identifies a  specific  instrument  within the Contractor
                laboratory.   No two ICP-MS instruments within a laboratory may
                have the same ICP-MS  Instrument  ID.

3.4.18.2.2     Report the  date   (formatted as MM/DD/YYYY) on  which  the ICP-MS
                tune was performed.   This  date shall not exceed the dates of
                analysis by ICP-MS in  the  Sample  Data Package.
 ILM05.2                              B-44

-------
                                                         Exhibit B -- Section 3
                                                              Form Instructions
                                                                      Form XV-IN

3.4.18.2.3     For "Avg. Measured Mass  (amu)", enter the average mass
               calculated  from  the  five tune  analyses ' (in  atomic mass units,
               to two decimals  places) measured for that isotope.

3.4.18.2.4     For "Avg. Peak Width at 5% Peak Height  (amu)" enter the average
               peak width  calculated  from the analysis  (in1 atomic mass units,
               to two decimal places) at 5% of the peak height.

3.4.18.2.5     For "%RSD", enter percent Relative Standard  Deviation of the
               absolute  signals (intensities) for each isotope  calculated from
               the five  tune analyses.

3.4.19   ICP-MS Internal Standards Relative Intensity Summary [Form XV-IN]

3.4.19.1    Purpose.  This form is used to report the relative  internal
            standard intensity  levels during  a run for ICP-MS.  The relative
            intensity of each of the  internal standards in  all  analyses
            performed by ICP-MS must be reported on the form.   If more than
            one ICP-MS instrument or  run is used, submit additional Form(s)
            XV-IN as appropriate.  All runs for the lowest  alphanumeric
            instrument must be  reported in ascending order  before proceeding
            to the runs  for the next highest  instrument.

3.4.19.2    Instructions.  Complete the header information  according to the
            instructions in Exhibit B, Section 3.3.  Complete the remainder of
            the form using the  following instructions.

3.4.19.2.1     For "ICP-MS Instrument ID", enter an identifier  that uniquely
               identifies  a specific  instrument within the  Contractor
               laboratory.  No  two  ICP-MS instruments within a  laboratory may
               have the  same ICP-MS Instrument ID.

3.4.19.2.2     For "Start  Date", enter the date (formatted  MM/DD/YYYY) on
               which the analysis run was started.

3.4.19.2.3     For "End Date",  enter the date (formatted MM/DD/YYYY) on which
               the analysis run was ended.

3.4.19.2.4     Under "EPA  Sample No.", enter  EPA sample number  of each
               analysis, including  all QC operations applicable to the SDG.
               All EPA  sample numbers must be listed in increasing
               chronological  (date  and  time)  order of  analysis, continuing to
               the next  Form XV for the instrument run, if  applicable.  The
               order must  agree with  the order reported on  Form XIII-IN for
               that run.   The analysis  date and time of other analyses not
               associated  with  the  SDG, but analyzed by the instrument in the
               reported  analytical  run, must  be reported.   Those analyses must
               be identified with EPA sample  number of "ZZZZZZ."  Samples
               identified  as "ZZZZZZ" need not have intensities reported  for
               internal  standards.

3.4.19.2.5     Under "Time", enter the time (in military format - HHMM) at
               which each  analysis  was performed.

3.4.19.2.6     Under "Internal  Standards %RI  for:", enter the chemical symbol
               and elemental expression number of the  internal  standard in the
               "Element" header field provided to indicate  the  internal
               standard  and elemental expression for which  the  Relative
               Intensity (RI) of the  internal standards will be calculated in
               that column.
                                      B-45                               ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Form DC-1

3.4.19.2.6.1      In the "Element" column, enter the internal  standard
                  relative intensity (to the  nearest whole number)  of the
                  internal standard  for each  sample analysis listed on  the
                  form (excluding "ZZZZZZ").   The  internal standard relative
                  intensity (%RI)  is calculated using  the following formula:

                  EQ.  15   Internal Standard Percent Relative Intensity

                                 % RI =  —'x 100
                                         O


                  WHERE,  "I0"  is  the intensity of  the  internal standard in  the
                  blank calibration  standard  and "In"  is  the intensity  of the
                  internal standard  in EPA sample  number in the same units.

3.4.19.2.7     Under the "Q" column  to  the right of each  "Element"  column,
               enter an "R" if the %RI  for a  field sample, PE, duplicate, or
               spike is less  than 60 or greater than  125; otherwise leave the
               field blank.

3.4.19.2.8     Columns of  internal standard RI must be entered left to  right
               starting with the internal standards of the lower mass on the
               first Form XV-IN and  proceeding to  the  following Form XV-IN as
               appropriate.  All Forms XV-IN  for the  lowest numeric instrument
               must be reported in ascending  order by  the run number before
               proceeding to the next Form XV.

3.4.19.3     All  field  samples  and all QC samples  (including calibration
             standards, ICVs,  CCVs, ICBs, CCBs, CRIs,  ICSs, LCS, PB, serial
             dilutions, duplicates, PE samples, and spikes) associated with the
             SDG  must be reported on  Form XV-IN.  The  run must be continuous
             and  inclusive of all analyses performed on the particular
             instrument during the run.

3.4.19.4     Submit  one Form XV-IN per run if  no more than 32 analyses,
             including  instrument calibration, were analyzed in the run.  If
             more than  32 analyses were performed in the run, submit additional
             Form(s) XV-IN as appropriate.  Each new run must be started on the
             first line of Form XV-IN.

3.5    Sample Log-In Sheet  [Form DC-1]

3.5.1     Purpose.  This form is used to document the  receipt and inspection of
          samples and  containers.  At least one original Form DC-1 is required
          for each sample shipping container  (e.g., cooler).  If the samples in
          a single sample shipping container must be assigned to more than one
          SDG,  the original Form DC-1 shall be placed with the deliverables  for
          the SDG of the lowest alpha-numeric  number and a copy of Form DC-1
          shall be placed with the deliverables for the other SDG(s).  The
          copies  should be identified as "copy(ies)",  and the location of the
          original should be noted on the copies.

3.5.2     Instructions

3.5.2.1     Sign and date the airbill.    (If an airbill is not received,
             include a hardcopy receipt requested from the shipping company or
             a printout of the shipping company's electronic tracking
             information).
 ILM05.2                               B-46

-------
                                                         Exhibit B — Section 3
                                                              Form Instructions
                                                              Form DC-1  (Con* t)

3.5.2.2     Examine the shipping container and record the presence/absence of
            custody seals and their condition  (i.e., intact, broken) in
            Item 1.

3.5.2.3     Record the custody seal numbers in Item 2.

3.5.2.4     Open the container, remove the enclosed sample documentation, and
            record the presence/absence of USEPA forms  (i.e., Traffic
            Reports/Chain of Custody Records, packing lists) and airbills or
            airbill stickers in Items 3 and 4.  Specify  if there is an airbill
            present or an airbill sticker in Item 4.  Record the airbill or
            sticker number in Item 5.

3.5.2.5     Remove the samples from the shipping container(s), examine the
            samples and the sample tags  (if present), and record the condition
            of the sample bottles  (i.e., intact, broken, leaking) and presence
            or absence of sample tags in Items 6 and 7.

3.5.2.6     Record the presence or absence of a cooler temperature indicator
            bottle in Item 8.

3.5.2.7     Record the cooler temperature in Item 9.

3.5.2.8     Review the sample shipping documents and compare the information
            recorded on all the documents and samples and mark the appropriate
            answer in Item 10.

3.5.2.9     The log-in date should be recorded at the top of Form DC-1; record
            the date and time of cooler receipt at the laboratory in Items 11
            and 12.

3.5.2.KLO    If there are no problems observed during receipt, sign and date
             (include the time) Form DC-1 and Traffic Report/Chain of Custody
            Record, and write the sample numbers in the  "EPA Sample No."
            column.

3.5.2.11    Record the pH for all aqueous samples received.

3.5.2.12    Record the appropriate sample tags and assigned laboratory
            numbers, if applicable.

3.5.2.13    Any comments should be made in the "Remarks" column.

3.5.2.14    Record the fraction designation (if appropriate) and the specific
            area designation  (e.g., refrigerator number) in the "Sample
            Transfer" block located in the bottom left corner of Form DC-1.
            Sign and date the sample transfer block.

3.5.2.15    For Items 1, 3, 4, 6, 7, 8 and 10, circle the appropriate
            response.  Responses can be underlined if this form is completed
            by automated equipment.  Unused columns and  spaces shall be
            crossed out, initialed, and dated.

3.5.2.16    If there are problems observed during receipt  (including samples
            that have not been preserved to the proper pH) or an answer marked
            with an asterisk  (e.g., "absent*") was circled, contact SMO and
            document the contact as well as resolution of the problem on a CLP
            Communication Log.  Following resolution, sign and date the forms
            as specified in the preceding paragraph and  note, where
            appropriate, the resolution of the problem.
                                      B-47                               ILM05.2

-------
Exhibit B — Section 3
Form Instructions
Form DC-2

3.6   Full Inorganics Complete SDG File  (CSF) Inventory Sheet  [Form DC-2]

3.6.1    Purpose.  The CSF  Inventory Sheet is used to record both the
         inventory of Complete SDG File  (CSF) documents and the  number of
         documents in the original Sample Data  Package which is  sent to  the
         USEPA Region.

3.6.2    Instructions

3.6.2.1     Organize all EPA-CSF documents as described in Exhibit B, Sections
            2 and 3.  Assemble the documents in Exhibit B, Section 2 in  the
            order specified on Form DC-2, and stamp  each  page  with the
            consecutive  number.   (Do not number Form DC-2).  Inventory the CSF
            by reviewing the document  numbers and recording page number  ranges
            in the columns  provided on Form  DC-2. The Contractor shall  verify
            and  record  in  the "Comments" section on  Form  DC-2  all intentional
            gaps in  the  page numbering sequence (for example,  "page  numbers
            not  used, XXXX-XXXX, XXXX-XXXX").   If there are no documents for a
            specific document type, enter an "NA" in the  empty space.

3.6.2.2     Certain  laboratory-specific  documents related to the CSF may not
            fit  into a  clearly defined category.  The laboratory should  review
            Form DC-2 to determine  if  it is  most appropriate to place  them
            under Categories 33,  34,  35, or  36.  Category 36 should  be used  if
            there is no appropriate previous category.  These  types  of
            documents should be  described or listed  in  the blanks under  each
            appropriate category.

3.6.2.3     If it is necessary to insert new or inadvertently  omitted
            documents,  the Contractor  shall  follow these  steps:

            ™-   Number  all documents  to be  inserted with  the  next sequential
                 numbers and file the  inserts in their logical  positions  within
                 the CSF (e.g.,  file document 1000 between documents  6 and 7).

            •Ť=   Identify where the  inserts  are  filed in the  CSF by recording
                 the document numbers  and their  locations  under the "Other
                 Records" section of Form DC-2  (e.g., document  1000 is filed
                 between 6  and 7).
 ILM05.2                              B-48

-------
                                                         Exhibit B -- Section 4
                                                            Data Reporting Forms
4.0   DATA REPORTING FORMS

      The data reporting forms are shown on the following pages.
                                      B-49                               ILM05.2

-------
THIS PAGE INTENTIONALLY LEFT BLANK

-------
  EXHIBIT B




INORGANIC FORMS
                                        ILM05.2

-------
                                  USEPA - CLP

                                   COVER PAGE

Lab Name:  	   Contract:
Lab Code: 	 Case No.:  	 NRAS No.:  	 SDG No.:

SOW No.:   	

                EPA  Sample  No.                          Lab Sample ID
                                                               ICP-AES  ICP-MS
 Were  ICP-AES  and  ICP-MS  interelement               (Yes/No)
 corrections applied?

 Were  ICP-AES  and  ICP-MS  background corrections     (Yes/No)
 applied?

     If yes, were raw data generated before          (Yes/No)
     application of background corrections?

 Comments:
 I  certify that this  data package  is  in  compliance  with  the  terms  and
 conditions of the contract,  both  technically and for  completeness,  for  other
 than the conditions  detailed above.   Release of  the data contained  in this
 hardcopy data package and in the  computer-readable data submitted on diskette
 (or via an alternate means of electronic  transmission,  if approved  in advance
 by USEPA)  has been authorized by  the Laboratory  Manager or  the  Manager's
 designee,  as verified by the following  signature.

 Signature:	 Name: 	
 Date:	              Title: 	
                                   COVER  PAGE                            ILM05.2

-------
                                  USEPA - CLP
                                     1A-IN
                         INORGANIC ANALYSIS DATA SHEET
                                              SPA SAMPLE NO.
Lab Name:

Lab Code:
                 Contract:
Case No.:
NRAS No.:
                                         SDG No.:
Matrix:  (soil/water)

Level: (low/med)  	

% Solids: 	
                       Lab Sample ID:

                       Date Received:
Concentration Units
                           or
                dry weight):
CAS No.
7429-90-5
7440-36-0
7440-38-2
7440-39-3
7440-41-7
7440-43-9
7440-70-2
7440-47-3
7440-48-4
7440-50-8
7439-89-6
7439-92-1
7439-95-4
7439-96-5
7439-97-6
7440-02-0
7440-09-7
7782-49-2
7440-22-4
7440-23-5
7440-28-0
7440-62-2
7440-66-6
57-12-5


Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide


Concentration


























C


























Q


























M


























Color Before:

Color After:

Comments:
        Clarity Before:

        Clarity After:
                  Texture:
                  Artifacts:
                                   FORM IA-IN
                                                     ILM05.2

-------
                                  USEPA - CLP
                                     1B-IN
                         INORGANIC ANALYSIS DATA SHEET
                                              EPA SAMPLE NO.
Lab Name:

Lab Code:
                 Contract:
Case No.:
NRAS No.:
                                         SDG No.:
Matrix:  (soil/water)

Level:  (low/med) 	

% Solids:    	
                        Lab  Sample  ID:

                        Date Received:
Concentration Units  ( BBMl or
                 dry weight):


Color Be
Color Af
Comments
CAS No.


























Analyte


























Concentration


























C


























Q


























M


























fore: Clarity Before: Texture:

ter: Clarity After: Artifacts:




                                    FORM IB-IN
                                                                         ILM05.2

-------
                                  USEPA - CLP

                                     2 A-IN
                INITIAL AND CONTINUING CALIBRATION VERIFICATION
Lab Name:

Lab Code:
                           Contract:
Case No.:
                   NRAS No.:
SDG No.:
Initial Calibration Verification Source:
Continuing Calibration Verification Source:

Concentration Units: ug/L

Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magne si-urn
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Initial Calibration
Verification
True




























Found




























%R(D




























Continuing Calibration Verification
True




























Found




























%R(D




























Found




























%R(D




























M




























 (1)  Control Limits: Mercury 80-120; Other Metals 90-110; Cyanide 85-115
                                  FORM IIA-IN
                                                     ILM05.2

-------
                                  USEPA - CLP

                                     2B-IN
                              CRQL CHECK STANDARD
Lab Name:

Lab Code:
        Contract:
                    Case  No.:
NRAS No.:
SDG No.
CRQL Check Standard Source:

Concentration Units: ug/L


Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide



CRQL Check Standard
Initial Final
True



























Found*



























%R(D



























Found*



























%R(D



























    (1)  Control Limits:  70-130 with the  following exceptions:
    ICP-AES - Antimony,  Lead,  and Thallium:  50-150.
    ICP-MS - Cobalt,  Manganese,  and Zinc:  50-150.

    *  If applicable,  enter the concentration qualifier "J"  or
 concentration in these  columns (e.g., 0.20U for Mercury).
                       "U" after the
                                   FORM IIB-IN
                                  ILM05.2

-------
                                  USEPA  - CLP

                                      3-IN
                                     BLANKS
Lab Name:

Lab Code:
                           Contract:
Case No.:
NRAS No.:
                                         SDG No.
Preparation Blank Matrix  (soil/water):
Preparation Blank Concentration Units  (ug/L or rag/kg):
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Initial
Calibration
Blank (ug/L)





























C




























Continuing Calibration
Blank (ug/L)
1




























C




























2




























C




























3




























C




























Preparation
Blank





























C





























M



























                                   FORM  III-IN
                                                                         ILM05.2

-------
                                  USEPA - CLP

                                     4A-IN
                       ICP-AES INTERFERENCE CHECK SAMPLE
Lab Name:

Lab Code:
                           Contract:
Case No.:
                   NRAS No.:
                                         SDG No.:
ICP-AES Instrument ID:
                                                       ICS Source:
Concentration Units: ug/L
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc




True
Sol. Sol.
A AB




















































Initial Found
Sol. Sol.
A %R .AB %R








































































































Final Found
Sol. Sol.
A %R AB %R








































































































                                   FORM IVA-IN
                                                     ILM05.2

-------
                                  USEPA - CLP

                                     43-IN
                        ICP-MS  INTERFERENCE CHECK SAMPLE
Lab Name:

Lab Code:
                            Contract:
Case No.:
                                       NRAS No.:
                                                            SDG No.:
ICP-MS Instrument ID:
                      ICS Source:
Concentration Units: ug/L
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Carbon
Chloride
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Molybdenum
Nickel
Phosphorus
Potassium
Selenium
Silver
Sodium
Sulfur
Thallium
Titanium
Vanadium
Zinc




True
Sol. Sol.
A AB
































































Found
Sol. Sol.
A %R AB %R
































































































































                                  FORM IVB-IN
                                                     ILM05.2

-------
                                  USEPA - CLP

                                     5A-IN
                          MATRIX SPIKE  SAMPLE RECOVERY
                                              EPA SAMPLE NO.
Lab Name:

Lab Code:
                 Contract,:
Case No.:
                   NRAS No.:
Matrix: (soil/water)

% Solids for Sample:
SDG No.:
                                    Level:  (low/med)
Concentration Units  (ug/L or mg/kg dry weight):
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Control
Limit
%R




























Spiked Sample
Result (SSR)
C
























































Sample
Result (SR)
C
























































Spike
Added (SA)




























%R




























Q




























M




























 Comments:
                                    FORM VA-IN
                                                      ILM05.2

-------
                                  USEPA - CLP
                                     5B-IN
                      POST-DIGESTION  SPIKE SAMPLE RECOVERY
                                              EPA SAMPLE NO.
Lab Name:

Lab Code:
                 Contract:
Case No.:
NRAS No.:
                                                             SDG No.:
Matrix: (soil/water)
                                    Level:  (low/med)
Concentration Units: ug/L
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide
v ^



Control
Limit
%R



















_
..






Spiked Sample
Result (SSR)
C






















































Sample
Result (SR)
C









.












































Spike
Added (SA)



























%R



























Q



























M



























Comments:
                                   FORM VB-IN
                                                     ILM05.2

-------
                                  USEPA - CLP
Lab Name:

Lab Code:
Case No.:
                                      6-IN
                                   DUPLICATES
Contract: _

  NRAS No.:
                                              EPA SAMPLE NO.
                                         SDG No.:
Matrix: (soil/water)

% Solids for Sample:
                                     Level:  (low/med)

                             % Solids for Duplicate:
Concentration Units  (ug/L or mg/kg dry weight):
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Control
Limit




























Sample (S)
C
























































Duplicate (D)
C



























.




























RPD




























Q




























M




























                                    FORM VI-IN
                                                      ILM05.2

-------
                                  USEPA - CLP

                                      7-IN
                           LABORATORY CONTROL SAMPLE
Lab Name:

Lab Code:
                            Contract:
Case No. :
                   NRAS No.:
SDG No.
Solid LCS Source:
Aqueous LCS Source:
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury '
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Aqueous (ug/L)
True Found %R




















































































Solid (mg/kg)
True Found C Limits %R















































•
























































































































                                   FORM VII-IN
                                                     ILM05.2

-------
                                  USEPA - CLP
                                      8-IN
                      ICP-AES and ICP-MS SERIAL DILUTIONS
                            EPA SAMPLE NO.
Lab Name:

Lab Code:
Contract:
                   Case No.
Matrix: (soil/water)
  NRAS No.:
SDG No.:
                 Level:  (low/med)
Concentration Units: ug/L
Analyte |
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc




Initial Sampl
Result (I)


























e
C


























Serial
Dilution
Result (S)


























C


























%
Difference





•




















Q


























M


























                                  FORM VIII-IN
                                    ILM05.2

-------
                                  USEPA - CLP

                                      9-IN
                      METHOD DETECTION LIMITS  (ANNUALLY)
Lab Name:

Lab Code:
                            Contract:
Case No.:
NRAS No.:
Instrument Type:
                                Instrument ID:
                      SDG  No.

                       Date:
Preparation Method:
Concentration Units  (ug/L or mg/kg):
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Wavelength
/Mass




























CRQL




























MDL




























Comments:
                                   FORM IX-IN
                                                                         ILM05.2

-------
                                  USEPA - CLP

                                     10A-IN
              ICP-AES INTERELEMENT CORRECTION FACTORS (QUARTERLY)
Lab Name:

Lab Code:
                            Contract:
Case No.:
ICP-AES Instrument ID:
NRAS No.:

  Date:
                                         SDG No.
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc




Wave-
length
(nm)


























Interelement Correction Factors for:
Al Ca Fe Mq


































































































































 Comments:
                                    FORM XA-IN
                                                      ILM05.2

-------
                                  USEPA - CLP

                                     10B-IN
              ICP-AES INTERELEMENT CORRECTION FACTORS (QUARTERLY)
Lab Name:

Lab Code:
                            Contract:
Case No.:
ICP-AES Instrument ID:
                   NRAS No.:

                      Date:
SDG No.
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc




Wave-
length
(nm)


























Interelement Correction Factors for:







•>


























































































































Comments:
                                   FORM XB-IN
                                                     ILM05.2

-------
                                  USEPA - CLP

                                      11-IN
                  ICP-AES and  ICP-MS  LINEAR  RANGES  (QUARTERLY)
Lab Name:

Lab Code:
                   Case No.:
ICP Instrument ID:
        Contract:
NRAS No.:

  Date:
SDG No.
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc




Integ.
Time
(Sec.)


























Concentration
(ug/L)


























M


























Comments:
                                    FORM XI-IN
                                  ILM05.2

-------
 USEPA  - CLP

     12-IN
PREPARATION LOG
Lab Name: Contract:
Lab Code: Case No.: NRAS No.: SDG No.:
Preparation Method:
EPA
Sample
No.



































Preparation
Date





>




























Weight
(gram)

































Volume
(mL)

































  FORM XII-IN
ILM05.2

-------
                                  USEPA - CLP

                                     13-IN
                               ANALYSIS RUN LOG
Lab Name:

Lab Code:
                           Contract:
Case No.:
NRAS No.:
                                         SDG No.
Instrument ID:

Start Date:
                         Analysis Method:

                         End Date:  	
EPA
Sample
No.





































D/F





































Time





































A
L





































S
B





































A
S





































B
A





































B
E





































C
D





































C
A





































C
R





































C
O





































C
U





































F
E




































Ar
P
B




































ial:
M
G




































/te.
M
N




































3
H
G





































N
I





































K





































S
E





































A
G





































N
A





































T
L





































V





































Z
N





































C
N
















































































































                                  FORM XIII-IN
                                                                         ILM05.2

-------
                                  USEPA - CLP

                                     14-IN
                                  ICP-MS Tune
Lab Name:

Lab Code:
Case No.:
ICP-MS Instrument ID:
                            Contract:
                   NRAS No.:
SDG No.

 Date:
Element - Mass
Be - 9
Mg - 24
Mg - 25
Mg - 26
Co - 59
In - 113
In - 115
Pb - 206
Pb - 207
Pb - 208




Avg. Measured Mass (amu)














Avg. Peak Width at
5% Peak Height (amu)














%RSD














Comments:
                                   FORM XIV-IN
                                                     ILM05.2

-------
                                  USEPA - CLP

                                     15-IN
             ICP-MS Internal Standards Relative Intensity Summary
Lab Name:

Lab Code:
Case No.:
ICP-MS Instrument ID:
                            Contract:
NRAS No.:
                      Start  Date:
SDG No.:  _

  End Date:
EPA Sample
No.
























Time























Internal Standards %RI For:
Element























Q























Element























Q























Element























Q























Element























Q























Element























Q























                                    FORM XV-IN
                                                      ILM05.2

-------
                                           SAMPLE LOG-IN SHEET
Lab Name
Received By (Print Name)
Page 	 of 	
Log-in Date
Received By (Signature)
Case Number
Remarks :
1. Custody Seal(s)
2. Custody Seal Nos.
Present /Absent*
Intact /Broken

3. Traffic Present/Absent*
Reports/Chain of
Custody Records or
Packing Lists
4. Airbill
5. Airbill No.
6. Sample Tags
Airbill/Sticker
Present /Absent*

Present /Absent*
Sample Tag Numbers Listed/Not
Listed on
Traffic
Report/Chain of
Custody Record
7 . Sample Condition
Intact/Broken*/
Leaking
8. Cooler Temperature Present/Absent*
Indicator Bottle
9. Cooler Temperature
10. Does information
on Traffic
Reports/Chain of
Custody Records
and sample tags
agree9
11. Date Received at
Lab
12 . Time Received
Yes/No*

Sample Transfer
Fraction
Area #
By
On
Fraction
Area f
By
On
Sample Delivery Group No.
EPA
Sample #


















Aqueous
Sample pH


















Corresponding
Sample Tag *


















Assigned
Lab 1


















NRAS Number

Remarks :
Condition of
Sample
Shipment, etc.


















* Contact SMO and attach  record of resolution
Reviewed By
Date


Logbook No.
Logbook Page No.
                                               FORM DC-1
ILM05.2

-------
                    FULL INORGANICS COMPLETE SDG FILE (CSF)  INVENTORY SHEET
    LABORATORY NAME

    CITY/STATE  	
     CASE  NO.  	 SDG NO.

     SDG NOs.  TO FOLLOW 	

     NRAS  NO.  	
     CONTRACT NO.

     SOW NO.  	
     All documents delivered in the Complete SDG File must be original documents
     where possible.   (Reference - Exhibit B Section 2.6)

                                                  PAGE NOs.               CHECK
                                               FROM         TO         LAB      REGION
1.   Inventory Sheet (DC-2)  (Do not number)

2.   Sample Log-In Sheet (DC-1)                 	        	        	      	

3.   Traffic Report/Chain of Custody Record     	        	        	      	

4.   Cover Page                                 	        	        	      	

5.   SDG Narrative                              	        ,	        	      	
    Inorganic Analysis
6.  Data Sheet (Form I-IN)

7.  Initial & Continuing Calibration
   - Verification  (Form IIA-IN)

8.  CRQL Standard
    (Form IIB-IN)

9.  Blanks  (Form III-IN)

10. ICP-AES Interference Check Sample
    (Form IVA-IN) '

11. ICP-MS Interference Check Sample
    (Form IVB-IN)

12. Matrix Spike Sample Recovery
    (Form VA-IN)

13. Post-Digestion Spike Sample Recovery
    (Form VB-IN)

14. Duplicates  (Form VI-IN)

15. Laboratory Control Sample
    (Form VII-IN)

16. ICP-AES and ICP-MS Serial Dilutions
    (Form VIII-IN)

17. Method Detection Limits  (Annually)
    (Form IX-IN)

18. ICP-AES Interelement Correction
    Factors  (Quarterly)  (Form XA-IN)

19. ICP-AES Interelement Correction
    Factors  (Quarterly)  (Form XB-IN)

20. ICP-AES and  ICP-MS Linear Ranges
    (Quarterly)  (Form XI-IN)

21. Preparation Log  (Form  XII-IN)

22. Analysis Run  Log  (Form XIII-IN)
                                           FORM DC-2-1                   .        ILM05.2

-------
                                                  PAGE NOs.                CHECK
                                               FROM         TO         LAB
23. ICP-MS Tune  (Form XIV-IN)

24. ICP-MS Internal Standards Relative
    Intensity Summary  (Form XV-IN)
25. ICP-AES Raw Data

26. GFAA Raw Data  (If Applicable)

27. ICP-MS Raw Data

28. Mercury Raw Data

29. Cyanide Raw Data

30. Preparation Logs Raw Data

31. Percent Solids Determination Log

32. USEPA Shipping/Receiving Documents
         Airbill (No.  of Shipments 	)

         Sample Tags

         Sample Log-In Sheet (Lab)    '

33. Misc. Shipping/Receiving Records
    (list all individual records)
         Telephone Logs
34. Internal Lab Sample Transfer Records &
    Tracking Sheets  (describe or list)
35. Internal Original Sample Prep &
    Analysis Records  (describe or list)
         Prep Records 	
         Analysis Records

         Description 	
36. Other Records  (describe or list)
         Telephone Communications Log
37. Comments:
Completed by:
(CLP Lab)  	
           (Signature)                        (Print Name & Title)    •          (Date)
Audited by:
(USEPA)    	   	   	
           (Signature)                        (Print Name & Title)              (Date)

                                           FORM DC-2-2                           ILM05.2

-------
THIS PAGE INTENTIONALLY LEFT BLANK

-------
          EXHIBIT C

INORGANIC TARGET ANALYTE LIST
   WITH  CONTRACT REQUIRED
     QUANTITATION LIMITS
             C-l                                ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               C-2

-------
           EXHIBIT C - INORGANIC TARGET ANALYTE LIST WITH CONTRACT
                          REQUIRED QUANTITATION  LIMITS

                               Table of Contents

Section                                                                     Page

1.0   INORGANIC TARGET ANALYTE LIST AND CONTRACT REQUIRED
      QUANTITATION LIMITS  (CRQLs )  	  5
                                      C-3                                ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               C-4

-------
                                                         Exhibit C — Section 1
                                        Inorganic Target Analyte List and CRQLs
1.0   INORGANIC TARGET ANALYTE LIST AND CONTRACT REQUIRED QUANTITATION LIMITS
      (CRQLs)
ICP-AES CRQL
Analyte CAS Number for Water1'2'3'4
(pg/L)
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide
7429-90-5
7440-36-0
7440-38-2
7440-39-3
7440-41-7
7440-43-9
7440-70-2
7440-47-3
7440-48-4
7440-50-8
7439-89-6
7439-92-1
7439-95-4
7439-96-5
7439-97-6
7440-02-0
7440-09-7
7782-49-2
7440-22-4
7440-23-5
7440-28-0
7440-62-2
7440-66-6
57-12-5
200
60
15
200
5
5
5000
10
50
25
100
10
5000
15
0.2
40
5000
35
10
5000
25
50
60
10
ICP-AES CRQL
for Soil1'2'3-"'5
(mg/kg)
40
12
3
40
1
1
1000
2
10
5
20
2
1000
3
0.1
8
1000
7
2
1000
5
10
12
1
ICP-MS CRQL
for Water1-2'4
(ug/L)
30
2
1
10
1
1
—
2
0.5
2
—
1
—
0.5
—
1
—
5
1
—
1
1
1
""
      1The CRQLs are the minimum levels of quantitation acceptable under the
contract Statement of Work  (SOW).

      2Subject to the restrictions specified in Exhibit D,  any analytical
method specified in ILM05.2 Exhibit D may be utilized as long as the
documented Method Detection Limits  (MDLs) are less than one-half the CRQLs.

      3Mercury is analyzed by cold vapor atomic absorption.  Cyanide is
analyzed by colorimetry/spectrophotometry.

      4Changes to the Inorganic Target Analyte List (TAL)   (e.g.,  adding an
additional analyte) or CRQLs may be requested under the flexibility clause in
the contract.

      5The CRQLs for soil are based on 100% solids and on the exact weights
and volumes specified in Exhibit D.  Samples with less than 100% solids may
have CRQLs greater than those listed in the table above.
                                       C-5
ILM05.2

-------
             EXHIBIT D




INTRODUCTION TO ANALYTICAL METHODS
          D-1/Introduction                         ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                         D-2/Introduction

-------
                        Exhibit D - Analytical Methods

                               Table of Contents

Section                                                                    Page

1.0   INTRODUCTION   	   5
      1.1   Inorganic Methods Flow Chart   	   5
      1.2   Figure 1 - Inorganic Methods Flow Char t	6
      1.3   Glassware Cleaning   	   6
      1.4   Standard Stock Solutions   	   6
      1.5   Verification of Aqueous Sample Preservatio n 	   6
      1.6   Percent Solids Determination Procedure   	   7
      1.7   Insufficient Sample Volume   	   8
      1.8   Sample Mixing 	   8
      1.9   Undiluted Analysis   	   8
      1.10  Dissolved Metals   	   9
      1.11  Replicate Exposure   	   9
      1.12  Raw Data Requirement s	9
      1.13  Quality Control Sample s  	   9
      1.14  Safety	•	9
      1.15  Pollution Prevention   	  10
      1.16  Waste Management   	  10


Part A -  Analytical Methods for Inductively Coupled Plasma - Atomic Emission
         Spectroscopy

Part B -  Analytical Methods for Inductively Coupled Plasma - Mass Spectrometry

Part C -  Analytical Methods for Cold Vapor Mercury Analysis

Part D -  Analytical Methods for Total Cyanide Analysis
                                D-3/Introduction                         ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                         D-4/Introduction

-------
                                                         Exhibit D — Section 1
                                                                   Introduction
1.0   INTRODUCTION
      The inorganic analytical service provides a contractual framework for
      laboratories.  This.framework applies USEPA Contract Laboratory Program
      (CLP) analytical methods for the isolation, detection, and quantitative
      measurement of 23 metals (including mercury) and cyanide in water/
      aqueous and/or soil/sediment samples.

      The analytical methods that follow are designed to analyze water and
      sediment/soil from hazardous waste sites for the presence of inorganic
      analytes contained on the Inorganic Target Analyte List (TAL)  (see
      Exhibit C).  The inorganic methods include alternative analysis
      procedures for some analytes, multiple preparation procedures, and
      Quality Control  (QC) requirements. Analytical techniques in the
      inorganic methodologies include Inductively Coupled Plasma - Atomic
      Emission Spectroscopy (ICP-AES), Inductively Coupled Plasma - Mass
      Spectroscopy  (ICP-MS), Cold Vapor Atomic Absorption Spectroscopy, and
      Spectrophotometry.  Graphite Furnace Atomic Absorption (GFAA) may be
      requested by the flexibility clause in the contract.

1.1   Inorganic Methods Flow Chart

      Figure 1 outlines the general analytical scheme the Contractor shall
      follow in performing standard trace metals and cyanide analyses under
      this contract.
                                D-5/Introduction                        ILM05.2

-------
Exhibit D — Section 1
Introduction (Con't)

1.2   Figure 1 - Inorganic Methods Flow Chart
                                 Field Sample
                            Traffic Report or SMO
                             Specified Parameters
                         Water/Aqueous  Soil/Sediment
                             Matrix        Matrix
                  Acid  Digestions
                     for Metals
                     Analysis
                     in Water
                Metals Analysis
               ICP-AES/ICP-MS/CVAA
      Acid Digestions
        for Metals
       Analysis in
      Soil/Sediment
        Metals Analysis
           ICP-AES/CVAA
                                  Data Reports
 1.3    Glassware Cleaning

       Lab glassware to be used within the metals analysis must be acid cleaned
       according to USEPA's manual,  Methods for Chemical Analysis of Water and
       Wastes or an equivalent procedure.  An electronic version can be found
       via USEPA's National Environmental Publications Internet Site (NEPIS)  at
       http://www.epa.gov/cincl.

 1.4    Standard Stock Solutions

       Stock solutions to be used for preparing instrument or method standards
       may be purchased or prepared as described in the individual methods of
       Exhibit D, Section "7 (Reagents and Standards) .

 1.5    Verification of Aqueous Sample Preservation

 1.5.1    At the time of sample receipt, the Contractor shall check the pH of
          the sample and note in a preparation log if the pH is less than 2 for
          metals.  In addition, it should be noted if the pH is greater than 12
          for a cyanide sample.  Unless instructed bV the USEPA Regional CLP
          Project Officer  (CLP PO), the Contractor shall not perform any pH
          adjustment action if the sample has not been properly preserved.  If
          the sample has not been properly preserved,  contact Sample Management
 ILM05.2
D-6/Introduction

-------
                                                         Exhibit D — Section 1
                                                           Introduction  (Con't)

         Office (SMO) for further instructions before proceeding with the
         preparation and analysis.  The Contractor may adjust the pH of a
         sample for metals if SMO provides written documentation to the
         Contractor from the USEPA Regional CLP PO or USEPA OERR Analytical
         Operations/Data Quality Center (AOC) Inorganic Program Manager  (AOC
         PM) authorizing the adjustment.

1.5.2    Before preparation is initiated for an aqueous cyanide sample,  the
         Contractor shall test for the presence of sulfides and oxidizing
         agents (e.g., residual chlorine).  The test for sulfides shall be
         performed by placing a drop of the sample on a strip of lead acetate
         paper  (which has been pre-moistened with pH 4 acetate buffer
         solution).  If the test strip turns black, the Contractor shall treat
         the total volume of sample with powdered cadmium carbonate or lead
         carbonate.  Yellow cadmium sulfide precipitates when the sample
         contains sulfide.  This operation shall be repeated until a drop of
         the treated sample solution does not darken the lead acetate test
         paper.  The solution shall be filtered through a dry filter paper
         into a dry beaker, and the volume of sample to be used for analysis
         shall be measured from the filtrate.  It is recommended that the
         Contractor avoid a large excess of cadmium carbonate and a long
         contact time in order to minimize a loss by complexation or occlusion
         of cyanide on the precipitated material.  The test for oxidizing
         agents shall be performed by placing a drop of the sample on a strip
         of potassium iodide - starch test paper  (KI - starch paper).  If the
         test strip turns blue, the Contractor shall contact SMO for further
         instructions from the Region before proceeding with sample
         preparation and analysis.  The Contractor shall document the presence
         of sulfides or oxidizing agents in the Sample Delivery Group (SDG)
         Narrative.

1.6   Percent Solids Determination Procedure

1.6.1    Immediately following the weighing of the sample to be processed for
         analysis, add 5-10 g of sample to a tared weighing dish.  Weigh and
         record the weight to the nearest 0.01 g.

1.6.2    Place weighing dish plus sample,  with the cover tipped to allow for
         moisture escape, in a drying oven maintained at 103-105°C.  Sample
         handling and drying should be conducted in a well-ventilated area.

1.6.3    Dry the sample overnight (12-24 hours) but no longer than 24 hours.
         If dried less than 12 hours, it must be documented that constant
         weight was attained.1  Remove the sample from the oven and cool in a
         desiccator with the weighing dish cover in place before weighing.
         Weigh and record weight to nearest 0.01 g.  Do not analyze the dried
         sample.

1.6.4    Duplicate percent solids determinations are required at the same
         frequency as other analytical determinations.  Duplicate results are
         to be recorded on Form VI-IN.
      *Drying time is defined as the elapsed time in the oven;  thus raw data
must record time in and out of the oven to document the 12-hour drying time
minimum.  In the event it is necessary to demonstrate the attainment of
constant weight, data must be recorded for a minimum of two repetitive
weigh/dry/desiccate/weigh cycles with a minimum of 1-hour drying time in each
cycle.  Constant weight would be defined as a loss in weight of no greater
than 0.01 g between the start weight and final weight of the last cycle.

                                D-7/Introduction                        ILM05.2

-------
Exhibit D — Section 1
Introduction (Con't)

1.6.5    For the duplicate percent solids determination, designate one sample
         aliquot as the "original" sample and the other aliquot as the
         "duplicate" sample.  Calculate dry weight using the results of the
         "original" sample aliquot.

1.6.6    Calculate percent solids by the formula below.  The value thus
         obtained will be reported on  the appropriate Forms I and, where
         applicable, Forms VA-IN and VI-IN.  This value will be used for
         calculating analytical concentration on a dry weight basis.

         EQ. 1  Percent Solids
                           , .  ,    Sample Dry Weight   ,__
                       %.Solids =	— x 100
                                  Sample Wet Weight
 1.6.7     If  the  sample  contains  less  than  50%  solids,  the Contractor  shall
          notify  SMO immediately  of  the samples impacted.  After notification
          to  SMO,  the Contractor  shall proceed  with sample analysis  and
          document the issue in the  SDG Narrative.

 1.7    Insufficient Sample Volume

       If  insufficient sample volume (less  than the required  amount) is
       received to perform the analysis,  the Contractor shall contact  the  SMO
       to  apprise them of the problem.   SMO will contract the Region for
       instructions.   The Region  will  either approve that no  sample  analysis be
       performed  or will require  that  a reduced volume  be used for the sample
       analysis.   No other changes in  the analysis  will be permitted.   SMO will
       notify the Contractor of the  Region's decision.   The Contractor shall
       document the Region's decision  in the SDG Narrative.

 1.8    Sample Mixing

       Unless instructed otherwise by  the USEPA Regional CLP  PO,  all samples
       shall  be mixed thoroughly  prior to aliquoting for digestion.  There is
       no  specific procedure provided  herein for homogenization of soil/
       sediment samples; however, an effort should be made to obtain a
       representative aliquot.

 1.9    Undiluted  Analysis

 1.9.1     All samples shall  be run undiluted for multi-analyte analysis (i.e.,
          the final product  of the  sample preparation procedure)  unless-the
          dilution adjusted  detection  limits for all analytes are below the
          CRQL.  When an analyte  concentration exceeds  the calibrated  or linear
          range,  appropriate dilution   (but  not below the CRQL) and re-analysis
          is  required.  The  Contractor shall use the least dilution  necessary
          to  bring the analyte(s) instrument reading within  the upper  half of
          the calibrated or  linear  range and report the highest valid  value for
          each analyte as measured  from the undiluted and diluted analyses.
          Unless the Contractor can  submit  proof that dilution was required to
          obtain valid results, both diluted and undiluted sample measurements
          must be contained in the  raw data.

 1.9.2     For single analyte analysis, a diluted sample analysis may be the
          only sample analysis performed if the analyte's  instrument result is
          in the upper half  of the  calibration range.  An  undiluted  sample
          analysis does not  have to  be performed in this case.  The  sample and
          its associated matrix spike  and duplicate shall  initially  be run at
          the same dilution.


 ILM05.2                         D-8/Introduction

-------
                                                         Exhibit D — Section 1
                                                           Introduction  (Con't)


1.9.3    All sample dilutions shall be made with reagent water appropriately
         acidified  (except for cyanide) to maintain constant acid strength.

1.10  Dissolved Metals

1.10.1   If dissolved metals are requested by USEPA Regional Offices, the
         Contractor shall follow the instructions provided on the Traffic
         Report(s)/Chain of Custody Record(s).  If there are no instructions
         on the Traffic Report/Chain of Custody Record, the Contractor shall
         digest the samples designated as dissolved metals.

1.10.2   If the Regional Office indicates on the Traffic Report/Chain of
         Custody Record that a digestion is not to be performed when analyzing
         field samples for dissolved metals, then a aqueous Laboratory Control
         Sample  (LCSW) and a post-digestion spike sample  (hardcopy Form VB-IN
         and diskette QC codes PDO and PDF) are not required.

1.11  Replicate Exposure

      If the Contractor analyzes samples using multiple injections or
      exposures, the Contractor must use the data obtained from all injections
      or exposures to calculate the final sample result even if more than the
      minimum number of injections or exposures are taken.

1.12  Raw Data Requirements

      The Contractor is reminded and cautioned that the collection and
      provision of raw data may or may not be referred to within the
      individual methods of Exhibit D or the Quality Assurance (QA) protocol
      of Exhibit E.  The raw data deliverable requirements are specified in
      Exhibit B, Section 2.5.2.3.  Raw data collected and provided in
      association with the performance of analyses under this contract shall
      conform to the appropriate provisions of Exhibit B.

1.13  Quality Control Samples

      If the Sampler designated two (or more) samples as QC for the same
      matrix, and the QC samples are not specifically labeled with the
      analysis they are to be used for (dissolved metals and total metals),
      then the Contractor is to contact SMO to report the issue.   SMO shall
      then contact the Region and notify the Contractor of the Regional
      decision.

1.14  Safety

      The toxicity or carcinogenicity of each reagent used in this SOW has not
      been precisely defined; however, each chemical compound should be
      treated as a potential health hazard.  From this viewpoint, exposure to
      these chemicals must be' reduced to the lowest possible level by whatever
      means available.  The laboratory is responsible for maintaining a
      current awareness file of Occupational Safety and Health Administration
       (OSHA) regulations regarding the safe handling of chemicals specified in
      this method.  A reference file of material handling data sheets should
      also be made available to all personnel involved in the chemical
      analysis.
                                D-9/Introduction                        ILM05.2

-------
Exhibit D — Section 1
Introduction (Con't)

1.15  Pollution Prevention

1.15.1   Pollution prevention encompasses any technique that reduces or
         eliminates the quantity or toxicity of waste at the point of
         generation.  Numerous opportunities for pollution prevention exist in
         laboratory operation.  USEPA has established a preferred hierarchy of
         environmental management  techniques that places pollution prevention
         as the management  option  of first choice.  Whenever feasible,
         laboratory personnel should use pollution prevention techniques to
         address  their waste generation.  When wastes cannot be  feasibly
         reduced  at the source, USEPA recommends recycling as the next best
         option.

1.15.2   For information about pollution prevention that may be  applicable to
         laboratories and research institutions consult "Less is Better:
         Laboratory Chemical Management for Waste Reduction," available from
         the American Chemical Society's Department of Government Relations
         and Science Policy, 1155  16th Street, N.W., Washington  D.C., 20036,
          (202)  872-4477.

1.16  Waste Management

      USEPA requires that laboratory waste management practices  be conducted
      consistent  with all applicable rules and regulations.  USEPA urges
      laboratories  to protect the  air, water, and land by minimizing and
      controlling all releases from hoods and bench operations,  complying with
      the  letter  and spirit of any sewer discharge permits and regulations,
      and  by complying with all  solid and hazardous waste regulations,
      particularly  the hazardous waste identification rules and  land disposal
      restrictions.  For  further information on waste management consult "The
      Waste Management Manual for  Laboratory Personnel", available from the
      American  Chemical Society  at the address listed in Section 1.15.2.
 ILM05.2                        D-10/Intr'oduction

-------
     EXHIBIT D - PART A

     ANALYTICAL METHODS
             FOR
INDUCTIVELY COUPLED PLASMA -
ATOMIC EMISSION SPECTROSCOPY
        D-1/ICP-AES                            ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                        D-2/ICP-AES

-------
                  Exhibit D - Analytical Methods  for  ICP-AES

                               Table of Contents

Section                                                                     Page


1.0   SCOPE AND.APPLICATION 	   5

2.0   SUMMARY OF METHOD	5

3.0   DEFINITIONS 	   5

4.0   INTERFERENCES 	   6
      4.1   Spectral Interferences   	  6
      4.2   Physical Interferences   	  6
      4.3   Chemical Interferences   	  6

5.0   SAFETY   	7

6.0   EQUIPMENT AND SUPPLIES   	  7
      6.1   Glassware/Labware 	   7
      6.2   Inductively Coupled Plasma - Atomic Emission Spectrometer
             (ICP-AES)  	  7

7.0   REAGENTS AND STANDARDS   	  8
      7.1   Reagents   	8
      7.2   Standards	8

8.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE    	  12
      8.1   Sample Collection and Preservation    	  12
      8.2   Procedures for Sample Storage  	  12
      8.3   Procedure  for Sample Digestate Storage    	  12
      8.4   Contract Required Holding Time   	  12

9.0   CALIBRATION AND STANDARDIZATION 	  13
      9.1   Instrument Operating Parameters  	  13
      9.2   Microwave Calibration Procedure  	  13
      9.3   Inductively Coupled Plasma - Atomic Emission Spectrometer
             (ICP-AES)  Instrument Calibration Procedure   	  14
      9.4   Initial Calibration Verification (ICV)    	  14
      9.5   Continuing Calibration Verification  (CCV  )  	  15
      9.6   Initial and Continuing Calibration Blank  (ICB/CCB)    	  15

10.0  PROCEDURE	16
      10.1  Sample Preparation   	  16
      10.2  Microwave Digestion Cleaning Procedure    	  21
      10.3  Sample Analysis	21

11.0  DATA ANALYSIS AND CALCULATIONS   	23
      11.1  Water/Aqueous Sample Calculation   	  23
      11.2  Soil Sample Calculation	23
      11.3  Adjusted Method Detection Limit  (MDL)/Adjusted Contract
            Required Quantitation Limit  (CRQL) Calculation    	  23

12.0  QUALITY CONTROL  (QC)   	25
      12.1  Initial Calibration Verification (ICV)    	  25
      12.2  Continuing Calibration Verification  (CCV  )  	  25
      12.3  Contract Required Quantitation Limit  (CRQL) Check
            Standard  (CRI)   	25
      12.4  Blank Analyses   	25
      12.5  Interference Check Sample (ICS)  	  27
      12.6  Spike Sample Analysi s	28
      12.7  Duplicate Sample Analysi s 	  29
      12.8  Laboratory Control Sample (LCS) Analysis    	  30


                                  D-3/ICP-AES                           ILM05.2

-------
                  Exhibit  D  -  Analytical Methods  for  ICP-AES

                           Table of Contents  (Con't)

Section                                                                     Page

      12.9  ICP-AES Serial Dilution Analysis   	  30
      12.10 Method Detection Limit  (MDL) Determination    	  31
      12.11 Interelement Corrections   	  32
      12.12 Linear Range Analysis Standard  (LRS)    	  32

13.0  METHOD PERFORMANCE	33

14.0  POLLUTION PREVENTION   	  33

15.0  WASTE MANAGEMENT   	  33

16.0  REFERENCES   	33

17.0  TABLES/DIAGRAMS/FLOWCHARTS   	  34
      TABLE 1:    Interferent  and Analyte Elemental Concentrations
                  Used for ICP-AES  Interference  Check Sample  (ICS)    ....  34
      TABLE 2:    Spiking  Levels for  Spike  Sample Analysis    	  35
 ILM05.2                        D-4/ICP-AES

-------
                                            Exhibit D (ICP-AES)  — Sections 1-3
                                                          Scope and Application
1.0   SCOPE AND APPLICATION
      The following method is an inductively coupled atomic plasma-atomic
      emission spectroscopy procedure that is used to analyze water, sediment,
      sludge, and soil samples taken from hazardous waste sites.  All metals
      (except mercury) which are contained in the Inorganic Target Analyte
      List  (TAL) in Exhibit C are quantitated by this Inductively Coupled
      Plasma - Atomic Emission Spectroscopy  (ICP-AES) method.

2.0   SUMMARY OF METHOD

      Water and soil samples are treated with acids and heat or microwave
      energy to solubilize the metals present.  These digestates are then
      analyzed for trace metals by an atomic emission optical spectroscopic
      technique.  Samples are nebulized and the aerosol that is produced is
      transported to a plasma torch where excitation occurs.  Characteristic
      atomic-line emission spectra are produced by a radio-frequency
      inductively coupled plasma.  The spectra are dispersed and the
      intensities of the lines are monitored by a photosensitive device.  The
      signals from the photosensitive device are processed by a computer.  A
      background correction technique is required to compensate for variable
      background contribution to the spectra of trace elements.  Background
      must be measured adjacent to analyte lines on samples during analysis.
      The position selected for the background intensity measurement, on
      either or both sides of the analytical line, will be determined by the
      complexity of the spectrum adjacent to the analyte line.  The position
      used must be free of spectral interference and reflect the same change
      in background intensity as occurs at the analyte wavelength measured.
      Background correction is not required in cases of line broadening where
      a background correction measurement would actually degrade the
      analytical result.

3.0   DEFINITIONS

      See Exhibit G for a complete list of definitions.
                                  D-5/ICP-AES
ILM05.2

-------
Exhibit D (ICP-AES) — Section 4
Interferences

4.0   INTERFERENCES

      Several types of interference effects may contribute to inaccuracies in
      the determination of trace elements in water and soil/sediments.  To
      prevent this, appropriate steps must be taken in all analyses to ensure
      that potential interferences are taken into account.  This is especially
      true when dissolved solids exceed 1500 milligrams per Liter  (mg/L) and
      when total elements are determined after the appropriate digestion
      procedures are performed.  Several types of interferences are summarized
      below:

4.1   Spectral Interferences

      Spectral interferences can be categorized as: overlap of a spectral line
      from another element, unresolved overlap of molecular band spectra,
      background contribution from continuous or recombination phenomena,
      and/or background contribution  from stray light from the line emission
      of high concentration elements.  The first of these effects  can be
      compensated  by utilizing  a computer correction of the raw data.  This
      would require the monitoring and measurement of the interfering element.
      The second effect may require selection of an alternate wavelength.  The
      third and fourth effects  can usually be compensated by a background
      correction adjacent to the analyte line.  In addition, users of
      simultaneous multi-element instrumentation must assume the
      responsibility of verifying the absence of spectral interference from an
      element that could occur  in a sample but for which there is  no channel
      in the instrument array.

 4.2   Physical Interferences

      Physical interferences are generally considered to be effects associated
      with the sample  nebulization and transport processes.  Such  properties
      as change in viscosity and surface tension can cause significant
      inaccuracies especially in samples which may contain high dissolved
      solids and/or acid concentrations.  The use of a peristaltic pump may
      minimize these  interferences.   If these types of interferences are
      present, they must be  reduced by dilution of the sample.

      Another problem  which  can occur from high dissolved solids is salt
      buildup at  the  tip of  the nebulizer.   This affects aerosol flow rate
      causing instrumental drift.  Wetting the argon prior to nebulization,
      the use of  a tip washer,  or sample dilution has been used to control
      this problem.  Also, it has been reported that better control of the
      argon  flow  rate  improves  instrument performance.  This is accomplished
      with the use of  mass flow controllers.

 4.3   Chemical  Interferences

      Chemical  interferences  are characterized by molecular compound
       formation,  ionization  effects and  solute vaporization effects.  Normally
       these  effects  are  not  pronounced with  the  Inductively Coupled Plasma -
      Atomic Emission  Spectrometer  (ICP-AES)  technique; however, if observed
       they  can be minimized  by  careful selection of operating conditions  (that
       is,  incident power,  observation position, and  so  forth), by  buffering of
       the  sample,  and by matrix matching.  These types of  interferences can be
       highly dependent on  matrix  type and  the  specific analyte element.
 ILM05.2                        D-6/ICP-AES

-------
                                          Exhibit D (ICP-AES) — Sections 5 & 6
                                                                          Safety

5.0   SAFETY

      See Section 1.14 in Exhibit D - Introduction to Analytical Methods.

6.0   EQUIPMENT AND SUPPLIES

      Brand names, suppliers, and part numbers are for illustrative purposes
      only.  No endorsement is implied.  Equivalent performance may be
      achieved using equipment and supplies other than those specified here,
      however, a demonstration of equivalent performance meeting the
      requirements of this Statement of Work  (SOW) is the responsibility of
      the Contractor.  The Contractor shall document any use of alternate
      equipment or supplies in the Sample Delivery Group  (SDG) Narrative.

6.1   Glassware/Labware

6.1.1    250 milliliter  (mL) beaker or other appropriate vessel

6.1.2    Watch glasses

6.1.3    Funnels

6.1.4    Graduated cylinders

6.1.5    Various volumetric flasks (Type A)

6.1.6    Thermometer that covers a range of 0-200°C

6.1.7    Whatman No. 42  filter paper or equivalent

6.1.8    Hot plate, block digester, or other heating source

6.1.9    Equipment and supplies for microwave digestion

6.1.9.1     Whatman No.  41 filter paper  (or equivalent)

6.1.9.2     Disposable polypropylene filter funnel

6.1.9.3     Polyethylene bottles, 125 mL, with caps

6.1.9.4     Microwave oven with programmable power settings  up to at least 600
            watts.

6.1.9.5     The system must use PTFE PFA digestion vessels  (120 mL capacity)
            capable of withstanding pressure of  up to 110  (ą10) pounds per
            square inch  (psi)  [7.5  (ą0.7 atm)].   These vessels are capable of
            controlled pressure relief at pressures exceeding 110 psi.

6.1.9.6     A rotating turntable must be used to  ensure homogeneous
            distribution of microwave radiation  within the  oven.  The speed  of
            the turntable must be a minimum of 3  revolutions per minute  (rpm).

6.1.10   Balances - Analytical Balance, 300 gram  (g) capacity, and minimum
         ą0.01 g.

6.2   Inductively Coupled Plasma - Atomic Emission Spectrometer (ICP-AES)
      consisting of a computer controlled atomic  emission spectrometer with
      background correction, a radio frequency generator, and a supply of
      Argon gas, welding grade or better.
                                  D-7/ICP-AES                            ILM05.2

-------
Exhibit D (ICP-AES) — Section 7
Reagents and Standards

7.0   REAGENTS AND STANDARDS

7.1   Reagents

      Acids used in the preparation of standards and for sample processing
      must be ultra-high purity grade or equivalent.   (Redistilled acids are
      acceptable.)

7.1.1    Reagent water - The purity of this water must be equivalent to ASTM
         Type II water  (ASTM D1193-77).  Use this preparation  for all
         reagents, standards,  and  dilutions of  solutions.

7.1.2    Acetic acid  - Concentrated  (specific gravity 1.06).

7.1.3    Hydrochloric acid  - Concentrated  (specific gravity  1.19).

7.1.4    Hydrochloric acid,  (1+1)  - Add 500 milliliters  (mL) cone. HC1
          (specific gravity  1.19)  to  400 mL reagent water and dilute  to  1 Liter
          (L) .

7.1.5    Nitric acid  - Concentrated  (specific gravity  1.41).

7.1.6    Nitric acid, (1+1) -  Add 500 mL cone.  HNO 3 (specific gravity 1.41) to
         400  mL reagent  water  and dilute to  1 L.

7.1.7    Hydrogen  peroxide  (30%)

7.1.8    Nitric acid, 5%  (v/v)  -  Add  50 mL cone. HNO3 to 500 mL reagent water;
         dilute to 1  L.

7.2    Standards

7.2.1    Introduction

         The  Contractor  must  provide all  standards  to  be used  with  this
         contract.  These standards  may be used only after they have been
         certified according  to the  procedure  in Exhibit E,  Section  8.0.   The
         Contractor  must be able to  verify that the  standards-are certified.
         Manufacturer's  certificates of analysis must  be  retained by the
         Contractor  and presented upon request.

7.2.2    Stock  Standard  Solutions

7.2.2.1     Stock standard solutions  may be  purchased  or prepared from  ultra
             high  purity grade chemicals  or metals.   All salts  must be dried
             for 1 hour at 105°C unless otherwise specified.

             (CAUTION : Many metal  salts are extremely toxic and may be fatal if
             swallowed.   Wash hands thoroughly after handling)   Typical  stock
             solution preparation procedures  follow.

7.2.2.2     Aluminum solution, stock (1 mL = 100 ug  Al)  -  Dissolve  0.100  grams
             (g)  of aluminum metal in an acid mixture of 4  mL of (1+1)  HC1 and
             1 mL of  cone. HN03 in a beaker.   Warm gently to effect solution.
             When solution is complete,  transfer quantitatively to a liter
             flask, add an additional 10 mL of  (1+1)  HC1 and dilute to 1000 mL
             with reagent water.

 7.2.2.3     Antimony solution, stock (1 mL = 100 pg  Sb)  -  Dissolve  0.2669  g
             K(SbO)C4H406 in reagent water,  add 10 mL (1+1)  HC1 and dilute to
             1000 mL  with reagent water.
 ILM05.2                        D-8/ICP-AES

-------
                                               Exhibit D  (ICP-AES)  —  Section  7
                                                 Reagents  and  Standards  (Con't)

7.2.2.4     Arsenic solution, stock  (1 mL =  100 pg As) -  Dissolve  0.1320  g of
            As2O3 in 100 mL of reagent water containing 0.4 g NaOH.  Acidify
            the  solution with 2 mL cone. HN03  and dilute to 1000 mL with
            reagent water.

7.2.2.5     Barium solution,  stock (1 mL = 100 pg Ba) - Dissolve 0.1516 g
            BaCl2 (dried at 250°C for 2 hours) in 10 mL reagent water  with 1
            mL  (1+1) HC1.  Add 10.0  mL  (1+1)  HC1 and dilute to 1000 mL with
            reagent water.

7.2.2.6     Beryllium solution, stock  (1 mL  = 100 ug Be)  - Do  not  dry.
            Dissolve 1.966 grams  (g) BeSO 4  =*• 4H20, in reagent  water, add  10.0
            mL cone. HN03 and dilute to 1000 mL with reagent water.

7.2.2.7     Cadmium solution, stock  (1 mL =  100 ug Cd) -  Dissolve  0.1142  g CdO
            in a minimum amount of  (1+1) HN03 .   Heat to increase rate of
            dissolution.  Add 10.0 mL cone.  HNO3  and dilute to 1000 mL with
            reagent water.

7.2.2.8     Calcium solution, stock  (1 mL =  100 ug Ca) -  Suspend 0.2498 g
            CaCO3 dried  at 180°C for 1 hour  before weighing in reagent water
            and  dissolve cautiously  with a minimum amount of (1+1)  HNO 3.  Add
            10.0 mL cone. HN03 and dilute to 1000 mL with reagent  water.

7.2.2.9     Chromium solution, stock (1 mL = 100 pg Cr) -  Dissolve 0.1923 g of
            Cr03 in reagent water.  When solution is complete  acidify  with 10
            mL cone. HNO3 and dilute to 1000 mL with reagent water.

7.2.2.10    Cobalt solution,  stock (1 mL = 100 pg Co) - Dissolve 0.1000 g of
            cobalt metal in a minimum amount of  (1+1) HNO 3.  Add  10.0 mL (1+1)
            HC1  and dilute to 1000 mL with reagent water.

7.2.2.11    Copper solution,  stock (1 mL = 100 pg Cu) - Dissolve 0.1252 g CuO
            in a minimum amount of  (1+1) HN03 .   Add 10.0 mL cone.  HN03 and
            dilute to 1000 mL with reagent water.

7.2.2.12    Iron solution, stock  (1 mL = 100 pg Fe) - Dissolve 0.1430  g Fe203
            in a warm mixture of 20  mL  (1+1)  HC1 and 2 mL of cone.  HNO 3.
            Cool, add an additional  5 mL of  cone. HNO 3 and dilute  to  1000 mL
            with reagent water.

7.2.2.13    Lead solution, stock  (1  mL = 100 pg Pb) - Dissolve 0.1599  g
            Pb(NO3)2 in a minimum amount of  (1+1)  HN03.  Add 10.0 mL of cone.
            HN03 and dilute to 1000 mL with  reagent water.

7.2.2.14    Magnesium solution, stock  (1 mL  = 100 pg Mg)  - Dissolve 0.1658 g
            MgO  in a minimum  amount  of  (1+1)  HNO 3.   Add 10.0 mL cone. HN03 and
            dilute to 1000 mL with reagent water.

7.2.2.15    Manganese solution, stock  (1 mL  = 100 pg Mn)  - Dissolve 0.1000 g
            of manganese metal in the acid mixture,  10 mL cone. HC1 and  1 mL
            cone. HN03,  and dilute to 1000 mL with reagent water.

7.2.2.16    Nickel solution,  stock (1 mL = 100 pg Ni) - Dissolve 0.1000 g of
            nickel metal in 10 mL hot cone.  HN03 ,  cool and dilute to 1000 mL
            with reagent water.

7.2.2.17    Potassium solution, stock  (1 mL  = 100 pg K) -  Dissolve  0.1907 g
            KC1, dried  at 110°C, in  reagent  water.  Dilute to  1000 mL.
                                  D-9/ICP-AES                            ILM05.2

-------
Exhibit D  (ICP-AES) — Section 7
Reagents and Standards (Con't)

7.2.2.18    Selenium solution, stock  (1 mL = 100 ug Se) - Do not dry.
            Dissolve 0.1727  g H2Se03  (actual assay 94.6%) in reagent water and
            dilute  to  1000 mL.

7.2.2.19    Silver  solution,  stock  (1 mL = 100 ug Ag)  - Dissolve 0.1575 g
            AgN03 in 100 mL  of reagent water and 10 mL cone. HNO 3.  Dilute to
            1000 mL with  reagent  water.

7.2.2.20    Sodium  solution,  stock  (1 mL = 100 ug Na)  - Dissolve 0.2542 g NaCl
            in  reagent water. Add  10.0 mL cone. HNO  3 and dilute to 1000 mL
            with reagent  water.

7.2.2.21    Thallium solution, stock  (1 mL = 100 ug Tl) - Dissolve  0.1303 g
            T1NO3 in reagent water.  Add 10.0 mL cone. HNO3  and dilute to 1000
            mL  with reagent  water.

7.2.2.22    Vanadium solution, stock  (1 mL = 100 ug V) - Dissolve  0.2297
            NH,V03 in a minimum amount of cone. HN03.   Heat  to  increase rate  of
            dissolution.   Add 10.0  mL cone.  HN03  and dilute to 1000 mL with
            reagent water.

7.2.2.23    Zinc solution, stock  (1 mL = 100 ug  Zn) - Dissolve 0.1245  g  ZnO  in
            a minimum  amount of  dilute  HNO 3.  Add 10.0 mL cone. HNO3 and
            dilute to  1000 mL with  reagent  water.

7.2.3     Secondary Dilution  Standards

7.2.3.1     Mixed  Secondary  Dilution  Standards

            Prepare mixed secondary dilution standard solutions by diluting
            the appropriate  volumes of stock standards with acidified reagent
            water  to  obtain  the  final volume.   Mixed secondary dilution
             standard  solutions may be purchased.   The purchased standards
             shall  meet the requirements  in  Section 7.2.1.

7.2.4     Working Standards

7.2.4.1     The calibration  blank is  prepared  by diluting  2 mL of  (1+1)  HNO3
             and 10 mL of (1+1) HC1 to 100  mL with reagent  water.  Prepare a
             sufficient quantity to be used to flush the system between
             standards  and samples.

7.2.4.2     Contract  Required Quantitation  Limit (CRQL)  Check  Standard (CRI)

             The concentration of the analytes in the CRI shall be at the
             respective CRQLs.  Information regarding the"CRI shall be reported
             on Form IIB-IN.

7.2.4.3      Interference Check Sample (ICS)  Solution

             The ICS consists of two solutions: Solution A  (ICSA) and Solution
             AB (ICSAB).   ICSA consists of the interferents and ICSAB consists
             of the analytes mixed with the interferents.

7.2.4.4      Method Detection Limit (MDL)  Solution

             The MDL solution shall be at a concentration of 3 to 5 times the
             expected MDL.
 ILM05.2                        D-10/ICP-AES

-------
                                               Exhibit D (ICP-AES) — Section 7
                                                 Reagents and Standards (Con't)

7.2.4.5     Mixed Calibration Standard Solutions

7.2.4.5.1      Prepare mixed calibration standard solutions by combining
               appropriate volumes of the stock solutions in volumetric flasks
                (see Sections 7.2.4.5.2 through 7.2.4.5.7).  Add 2 mL of  (1+1)
               HNO3 and 10 mL of (1+1)  HC1 and dilute to 100 mL with reagent
               water  (see Note  in Section 7.2.4.5.6).  Prior to preparing the
               mixed  standards, each stock solution  should be analyzed
               separately to determine possible spectral interference or the
               presence of impurities.  Care should  be taken when preparing
               the mixed standards that the elements are compatible and
               stable.  Transfer the mixed standard  solutions to a FEP
               fluorocarbon or  unused polyethylene bottle for storage.  Fresh
               mixed  standards  should be prepared as needed with the
               realization that concentration can change with aging.  Although
               not specifically required, some typical calibration standard
               combinations follow.

7.2.4.5.2      Mixed  standard solution I - manganese, beryllium, cadmium,
               lead,  and zinc.

7.2.4.5.3      Mixed  standard solution II - barium,  copper, iron, vanadium,
               and cobalt.

7.2.4.5.4      Mixed  standard solution III - arsenic and selenium.

7.2.4.5.5      Mixed  standard solution IV - calcium,  sodium, potassium,
               aluminum, chromium, and nickel.

7.2.4.5.6      Mixed  standard solution V - antimony,  magnesium, silver and
               thallium.

               NOTE:  If the addition of silver to the recommended acid
               combination results in an initial precipitation, add 15 mL of
               reagent water and warm the flask until the solution clears.
               Cool and dilute  to 100 mL with reagent water.  For this acid
               combination, the silver concentration should be limited to 2
               milligrams per Liter  (mg/L).  Silver  under these conditions is
               stable in a tap  water matrix for 30 days.  Higher
               concentrations of silver require additional HC1.

7.2.4.5.7      Protect all standards from light.  Samples, sample digestates,
               and standards must be stored separately.
                                 D-11/ICP-AES                           ILM05.2

-------
Exhibit D (ICP-AES) — Section 8
Sample Collection, Preservation, and Storage

8.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE

8.1   Sample Collection and Preservation

      All samples  must be collected in glass or polyethylene containers.
      Water/aqueous samples must be preserved with nitric acid to pH less than
      2 immediately after collection.  All samples must be  iced or
      refrigerated at 4°C  (ą2°C) from the time of collection until digestion.

8.1.1    Dissolved Metals

         For the  determination  of  dissolved metals,  the sample must be
         filtered through a  0.45 micrometer  (urn) pore  diameter membrane  filter
         at the time of collection or as  soon as possible.  Use a portion of
         the sample to  rinse  the filter flask, discard this portion, and
         collect  the required volume of filtrate.  Preserve the filtrate with
         nitric acid to pH less than 2 immediately after  filtration.

8.2   Procedures  for Sample Storage

      The samples must be protected from  light and refrigerated at 4°C  (+2°C)
      from the time of  receipt  until  60 days after delivery of a complete,
      reconciled  data package to USEPA.   After 60 days the  samples may be
      disposed of in a manner that complies with all applicable regulations.

8.3   Procedure for Sample Digestate Storage

      Sample digestates must  be stored until 365 days  after delivery of  a
      complete, reconciled data package to USEPA.

8.4   Contract Required Holding Time

      The maximum holding  time  for metals is 180 days  from  Validated Time of
      Sample Receipt  (VTSR).
 ILM05.2                       D-12/ICP-AES

-------
                                               Exhibit D (ICP-AES)  — Section 9
                                                Calibration and Standardization

9.0   CALIBRATION AND STANDARDIZATION

9.1   Instrument Operating Parameters

      Because of the differences between various makes and models of
      satisfactory instruments, no detailed operating instructions can be
      provided.  Instead, the analyst should follow the instructions provided
      by the manufacturer of the particular instrument.  The Method Detection
      Limit (MDL), precision, linear dynamic range, and interference effects
      must be investigated and established for each individual analyte line on
      that particular instrument.  All measurements must be within the
      instrument linear range where correction factors are valid.  It is the
      responsibility of the analyst to verify that the instrument
      configuration and operating conditions used satisfy the analytical
      requirements and to maintain Quality Control (QC) data confirming
      instrument performance and analytical results.

9.2   Microwave Calibration Procedure

9.2.1    The calibration procedure is a critical step prior to the use of any
         microwave unit.  The microwave unit must be calibrated every six
         months.  The data for each calibration must be available for review
         during on-site audits.  In order that absolute power settings may be
         interchanged from one microwave unit to another, the actual delivered
         power must be determined.

9.2.2    Calibration of a laboratory microwave unit depends on the type of
         electronic system used by the manufacturer.   If the unit has a
         precise and accurate linear relationship between the output power and
         the scale used in controlling the microwave unit, then the
         calibration can be a two-point calibration at maximum and 40% power.
         If the unit is not accurate or precise for some portion of the
         controlling scale, then a multiple-point calibration is necessary.
         If the unit power calibration needs a multiple-point calibration,
         then the point where linearity begins must be identified.  For
         example: a calibration at 100, 99, 98, 97, 95, 90, 80, 70, 60, 50,
         and 40% power settings can be applied and the data plotted.  The non-
         linear portion of the calibration curve can be excluded or restricted
         in use.  Each percent is equivalent to approximately 5.5-6 watts and
         becomes the smallest unit of power that can be controlled.  If 20-40
         watts are contained from 99-100%, that portion of the microwave
         calibration is not controllable by 3-7 times that of the linear
         portion of the control scale and will prevent duplication of precise
         power conditions specified in that portion of the power scale.

9.2.3    The power available for heating is evaluated so that the absolute
         power setting  (watts) may be compared from one microwave to another.
         This is accomplished by measuring the temperature rise in 1 kilogram
         (kg) of water exposed to microwave radiation for a fixed period of
         time.  The water is placed in a PTFE beaker  (or a beaker that is made
         of some other material that does not absorb microwave energy) and
         stirred before measuring the temperature.  Glass beakers absorb
         microwave energy and may not be used.  The initial temperature of the
         water must be between 19 and 25°C.  The beaker is circulated
         continuously through the field for at least two minutes at full
         power.  The beaker is removed from the microwave, the water is
         stirred vigorously, and the final temperature is recorded.  The final
         reading is the maximum temperature reading after each energy
         exposure.  These measurements must be accurate to ą0.1°C and made
         within 30 seconds of the end of heating.  If more measurements are
         needed, do not use the same water until it has cooled down to room
         temperature.  Otherwise, use a fresh water sample.


                                 D-13/ICP-AES                           ILM05.2

-------
Exhibit D (ICP-AES) — Section 9
Calibration and Standardization  (Con't)

         The absorbed power  is determined by the  following formula:

         EQ. 1  Absorbed  Power

                                  (K)  (CJ (m) (DT)
                              p = 	Ł	
          WHERE,   P    =   The  apparent power absorbed by  the  sample  in watts
                           (joules per second).

                  K    =   The  conversion factor for thermochemical calories
                           per  second to  watts  (=4.184).

                  Cp   =   The  heat capacity,  thermal capacity,  or specific
                           heat (cal. g'1  "C'1) of water  (=1.0).

                  m    =   The  mass of the sample in grams (g).

                  DT   =   The  final temperature minus  the initial temperature
                           <°C).

                  t    =   The  time in seconds  (s) .

          Using 2 minutes and 1 kg of reagent water, the calibration equation
          simplifies to:


                                 P =  (DT)  (34.87)
          The microwave user can now relate power in watts to the percent power
          setting of the microwave.

 9.3    Inductively Coupled Plasma - Atomic Emission Spectrometer (ICP-AES)
       Instrument Calibration Procedure

 9.3.1     Instruments shall be calibrated daily or once every 24 hours and each
          time the instrument is set up.   The instrument standardization date
          and time shall be included in the raw data.

 9.3.2     The calibration standards shall be prepared as in Section 7.2.4.5.

 9.3.3     Calibrate the ICP-AES instruments according to instrument
          manufacturer's recommended procedures.  At least two standards shall
          be used for ICP-AES calibration.  One of the standards shall be a
          blank.

 9.3.4     Any changes or corrections to the analytical system shall be followed
          by recalibration.

 9.4    Initial Calibration Verification (ICV)

 9.4.1     Immediately after each of the ICP-AES systems have been calibrated,
          the accuracy of the initial calibration shall be verified and
          documented for every analyte by the analysis of the ICV solution (s)
          at each wavelength used.

 9.4.2     Only if the ICV solution(s) is (are) not available from USEPA, or
          where a certified solution of an analyte is not available from any
          source, analyses shall be conducted on an independent standard at a


 ILM05.2                        D-14/ICP-AES

-------
                                               Exhibit D (ICP-AES)  — Section 9
                                        Calibration and Standardization (Con't)

         concentration other than that used for instrument calibration, but
         within the calibration range.  An independent standard is defined as
         a standard composed of the analytes from a different source than
         those used in the standards for the instrument calibration.

9.4.3    The ICV solution(s)  shall be run at each wavelength used for
         analysis.  The values for the ICV shall be reported on Form IIA-IN.

9.5   Continuing Calibration Verification (CCV)

9.5.1    To ensure calibration accuracy during each analysis run, one of the
         following standards is to be used for the CCV and shall be analyzed
         and reported for every wavelength used for the analysis of each
         analyte, at a frequency of 10% or every 2 hours during an analysis
         run, whichever is more frequent.  The standard shall also be analyzed
         and reported for every wavelength used for analysis at the beginning
         of the run and after the last analytical sample.  The analyte
         concentrations in the CCV standard shall be different than the
         concentration used for the ICV and shall be one of the following
         solutions at or near one-half of the calibration standard:

         ^     USEPA Solutions

         •"     NIST Standards

         '"     A Contractor-prepared  standard  solution

         The same CCV standard shall be used throughout the analysis runs for
         a Sample Delivery Group  (SDG) of samples received.

9.5.2    Each CCV analyzed shall reflect the conditions of analysis of all
         associated analytical samples  (the preceding 10 analytical samples or
         the preceding analytical samples up to the previous CCV).  The
         duration of analysis, rinses, and other related operations that may
         affect the CCV measured result may not be applied to the CCV to a
         greater extent than the extent applied to the associated analytical
         samples.  For instance, the difference in time between a CCV analysis
         and the blank immediately following it, as well as the difference in
         time between the CCV and the analytical sample immediately preceding
         it, may not exceed the lowest difference in time between any two
         consecutive analytical samples associated with the CCV.

9.5.3    Information regarding the CCV shall be reported on Form IIA-IN.

9.6   Initial and Continuing Calibration Blank  (ICB/CCB)

      A calibration blank shall be analyzed at each wavelength used for
      analysis immediately after every ICV and CCV, at a frequency of 10% or
      every 2 hours during the run, whichever is more frequent.  The blank
      shall be analyzed at the beginning of the run and after the last
      analytical sample.

      NOTE: A CCB shall be analyzed immediately after the last CCV, and the
      last CCV shall be analyzed immediately after the last analytical sample
      of the run.  The results for the calibration blanks shall be reported on
      Form III-IN.
                                 D-15/ICP-AES                           ILM05.2

-------
Exhibit D  (ICP-AES) — Section 10
Procedure

10.0  PROCEDURE

10.1  Sample Preparation

10.1.1   If insufficient  sample amount  (less than  90% of the required amount)
         is received  to perform the  analyses,  the  Contractor shall  contact  the
         Sample Management  Office  (SMO)  to  inform  them  of  the problem.   SMO
         will  contact the Region for instructions.   The Region  will either
         require  that no  sample analyses be performed or will require that  a
         reduced  volume be  used for  the  sample analysis.   No other  changes  in
         the analyses will  be permitted.  The  Contractor shall  document  the
         Region's decision  in the  Sample Delivery  Group (SDG) Narrative.

10.1.2   If multiphase samples  (e.g., two-phase liquid  sample,  oily
         sludge/sandy soil  sample) are  received by the  Contractor,  the
         Contractor shall contact  SMO to apprise them of the type of sample
         received.   SMO will contact the Region.  If all phases of  the  sample
         are amenable to  analysis, the  Region  may  require  the Contractor to do
         any of the following:

         30    Mix the sample  and  analyze an aliquot from  the homogenized
                sample.

         30     Separate  the phases of the sample and analyze  one or more of
                the phases,  separately.   SMO will provide EPA  sample numbers
                for the additional  phases, if required.

         00     Do not analyze  the  sample.

10.1.2.1    If all of the phases are not amenable  to analysis  (i.e.,  outside
             scope), the  Region  may require  the Contractor  to  do any of  the
             following:

             :x:     Separate the phases and analyze the phase (s)  that is (are)
                   amenable to analysis.  SMO will provide  EPA sample numbers
                   for the additional phases,  if required.

             ••*>     DO not analyze the sample.

10.1.2.2    No other changes in the  analyses  will  be permitted.  The
             Contractor shall document the Region's decision in  the SDG
             Narrative.

10.1.3   Water/Aqueous Sample  Preparation

10.1.3.1     Preparation  Method/Code  (HW1)  (USEPA Method 200.7,  December 1982)

             Shake sample and transfer 50-100 milliliter (mL)  of well-mixed
             sample to a 250 mL heating vessel, add 2 mL of (1+1) HNO 3  and 10
             mL of  (1+1)  HC1 to the sample.   Cover with watch glass or similar
             cover and heat on a hot plate,  block digester, or equivalent
             heating source which is adjustable and capable of maintaining a
             temperature of 92-95°C for 2 hours or until sample volume is
             reduced to between 25 and 50 mL,  making certain sample does not
             boil.  Cool sample and filter to remove insoluble material.

             NOTE: In place of filtering, the sample, after dilution and
             mixing, may be centrifuged or allowed to settle by gravity
             overnight to remove insoluble material.

            'Adjust sample volume to 50-100 mL with reagent water.  The sample
             is now ready for analysis.   Concentrations so determined shall be


 ILM05.2                       D-16/ICP-AES

-------
                                              Exhibit D (ICP-AES)  — Section 10
                                                              Procedure  (Con't)

            reported as "total".  If volumes less than 100 mL are used, all
            other reagents shall be reduced appropriately  (e.g., if 50 mL is
            used, reduce reagent volumes by one-half).  The final volume of
            the digestate must equal the initial volume of the sample aliquot.

10.1.3.2    Preparation Method/Code (MW1)  (USEPA SW-846 Method 3015)

10.1.3.2.1     A 45 mL aliquot of the sample is measured into PTFE digestion
               vessels.

10.1.3.2.2     5 mL of high purity concentrated HN03 is added to the digestion
               vessels.

10.1.3.2.3     The caps with the pressure release valves are placed on the
               vessels hand tight and then tightened,  using constant torque,
               to 12 ft/lbs.  The weight of each vessel is recorded to
               0.02 gram  (g).

10.1.3.2.4     Place 5 sample vessels in the carousel, evenly spaced around
               its periphery in the microwave unit.   Venting tubes connect
               each sample vessel with a collection vessel.  Each sample
               vessel is attached to a clean, double-ported overflow vessel to
               collect any sample expelled from the sample vessel in the event
               of over pressurization.  Assembly of the vessels into the
               carousel may be done inside or outside the microwave.

10.1.3.2.5     This procedure is energy balanced for  five 45 mL water samples
                (each with 5 mL of acid)  to produce  consistent conditions.
               When fewer than five samples are digested, the remaining
               vessels must be filled with 45 mL of tap, deionized, or reagent
               water and 5 mL of concentrated nitric  acid.

10.1.3.2.6     Newer microwave ovens may be capable of higher power settings
               which may allow a larger  number of samples.  If the analyst
               wishes to digest more than  5 samples at a time, the analyst may
               use different power settings as long as they result in the same
               time temperature conditions defined  in the power programming
               for this method.

10.1.3.2.7     The initial temperature of  the samples  should be 24°C (ą1°C).
               The Preparation Blank  (PB)  must have 45 mL of deionized water
               and the same amount  (5 mL)  of acid that is added to the
               samples.

10.1.3.2.8     The microwave unit first-stage program must be set to give 545
               watts for 10 minutes and  the second-stage program to give 344
               watts for 10 minutes.  This sequence brings the samples to
               160°C  (ą4°C) in 10 minutes  and permits a slow rise to 165-170°C
               during the second 10 minutes.

10.1.3.2.9     Following the 20 minute program, the samples are left to cool
               in the microwave unit for 5 minutes, with the exhaust fan on.
               The samples and/or carousel may then be removed from the
               microwave unit.  Before opening the  vessels, let cool until
               they are no longer hot to the touch.

10.1.3.2.10    After the sample vessel has cooled, weigh the sample vessel and
               compare to the initial weight as reported on the preparation
               log.  Any sample vessel exhibiting a less than or equal to
               0.5 g loss into the overflow vessel  must have any excess  sample
               from the associated collection vessel  added to the original
               sample vessel before proceeding with the sample preparation.


                                 D-17/ICP-AES                            ILM05.2

-------
Exhibit D (ICP-AES) — Section 10
Procedure (Con't)

               Any sample vessel  exhibiting  a  greater  than  0.5  g  loss must be
               identified in the  preparation log  and the  sample redigested.

10.1.3.2.11    Sample Filtration  - The digested samples are shaken well to mix
               in  any condensate  within  the  digestion  vessel  before  being
               opened.   The digestates are then  filtered  into 50  mL  glass
               volumetric flasks  through Whatman  No. 41  (or equivalent)  filter
               paper and diluted  to  50 mL (if  necessary).   The  samples  are now
               ready for analysis.   The  sample results must be  corrected by  a
               factor of 1.11 in  order to report  final concentration values
               based on  an  initial volume of 45 mL.  Concentrations  so
               determined shall be reported  as "total".

10.1.3.3     Preparation  Method/Code  (MW2)  (ASTM Standard  D4309-91)

10.1.3.3.1     A 50 mL aliquot of the sample is measured  into PTFE digestion
               vessels.

10.1.3.3.2     3mLof high purity concentrated HN03 and 2 mL of concentrated
               HC1 is added to the digestion vessels.

10.1.3.3.3     Proceed as in Preparation Method/Code "MW1", Sections
                10.1.3.2.3 through 10.1.3.2.11.

10.1.3.3.4     Sample Filtration  - The digested  samples are shaken well to mix
                in  any  condensate  within  the  digestion  vessel  before  being
                opened.   If  necessary,  the digestates are  then filtered through
                filter paper and diluted  to  55  mL.  The samples  are now ready
                for analysis.  The sample results must  be  corrected by  a factor
                of  1.1  in order to report final concentration  values  based on
                an  initial volume  of  50 mL.   Concentrations so determined shall
                be  reported  as "total".

10.1.4    Soil/Sediment Sample Preparation

10.1.4.1     Preparation  Method/Code  (HS1) (USEPA Method 200.7,  December 1982)

10.1.4.1.1     Mix the sample thoroughly to  achieve homogeneity.  For  each
                digestion procedure,  weigh (to the nearest 0.01  g) a  1.0 to
                1.5 g portion of sample and transfer to a beaker.

10.1.4.1.2     Add 10 mL of 1:1 nitric acidJ (HNO 3), mix  the  slurry,  and  cover
                with a watch glass.   Heat the sample to 92-95°C on hot plate or
                block digester, and reflux for 10 minutes without boiling.
                Allow the sample to cool, add 5 mL of concentrated HNO3,
                replace the watch glass,   as appropriate,  and reflux for 30
                minutes.   Do not allow the volume to be reduced to less than 5
                mL while maintaining a covering of solution over the  bottom of
                the heating vessel.

 10.1.4.1.3      After the second reflux step has  been completed and the sample
                has cooled,  add 2 mL of reagent water and 3 mL of 30% hydrogen
                peroxide (H202) .   Return  the  heating vessel  to the heat  source
                for warming to start the  peroxide reaction.  Care must be taken
                to  ensure that losses do  not occur due to excessively vigorous
                effervescence.  Heat until effervescence subsides, and cool the
                heating vessel.
 ILM05.2                        D-18/ICP-AES

-------
                                              Exhibit D (ICP-AES) — Section 10
                                                              Procedure  (Con't)

               Continue to add 30% H202  in  1  mL  aliquots with warming until
               the effervescence is minimal  or  until the general  sample
               appearance is unchanged.
                                                                            )
               NOTE: Do not add more than  a  total of 10 mL  30% H2 02.

10.1.4.1.4     Add 5 mL of 1:1 HC1 and 10 mL of reagent water, return the
               covered heating vessel to the heat source, and heat  for an
               additional 10 minutes.  After cooling,  filter through Whatman
               No. 42 filter paper  (or equivalent) and dilute to  100 mL  with
               reagent water.

               NOTE: In place of filtering,  the sample (after dilution and
               mixing) may be centrifuged  or allowed to settle by gravity
               overnight to remove insoluble material.

               Dilute the digestate 1:1  (200 mL final  volume) with  acidified
               water  [prepared by diluting 2 mL of  (1+1) HN03 and 10 mL of
                (1+1) HC1 to 100 mL] to maintain constant acid strength.  The
               sample is now ready for analysis.

10.1.4.2    Preparation Method/Code (HS2)  (USEPA SW-846 Method 3050B)

10.1.4.2.1     Mix the sample thoroughly to  achieve homogeneity.  For each
               digestion procedure, weigh  (to the nearest 0.01 g) a 1.0  to 2.0
               g portion of sample and transfer to a beaker.

10.1.4.2.2     Add 10 mL of 1:1 nitric acid  (HNO 3) ,  mix the slurry, and cover
               with a watch glass.  Heat the sample to 92-95°C on hot plate,
               block digester, or equivalent heating source,  and reflux  for 10
               minutes without boiling.  Allow  the sample to cool,  add 5 mL of
               concentrated HN03, replace the watch glass,  as appropriate,  and
               reflux for 30 minutes.  Do  not allow the volume to be reduced
               to  less than 5 mL while maintaining a covering of  solution  over
               the bottom of the heating vessel.  Add  an additional 5 mL of
               concentrated HN03 and reflux.  Repeat this step until sample
               oxidation is complete  (no brown  fumes generated).

10.1.4.2.3     After the reflux steps have been completed and the sample has
               cooled, add 2 mL of reagent water and 3 mL of 30%  hydrogen
               peroxide  (H202).   Return the heating vessel  to the heat source
               for warming to start the peroxide reaction.   Care  must be taken
               to  ensure that losses do not  occur due  to excessively vigorous
               effervescence.  Heat until  effervescence subsides, and cool the
               heating vessel.

10.1.4.2.4     Continue to add 30% H2O2 in  1  mL  aliquots with warming until
               the effervescence is minimal  or  until the general  sample
               appearance is unchanged.

               NOTE: Do not add more than  a  total of 10 mL  30% H2 02.

10.1.4.2.5     Add 10 mL of concentrated HC1 and return the  covered heating
               vessel to the heat source and heat for  an additional 10
               minutes.  After cooling,  filter  through Whatman No.  42 filter
               paper  (or equivalent) and dilute to 100 mL with reagent water.

               NOTE: In place of filtering,  the sample (after dilution and
               mixing) may be centrifuged  or allowed to settle by gravity
               overnight to remove insoluble material.

               The sample is now ready for analysis.


                                 D-19/ICP-AES                            ILM05.2

-------
Exhibit D  (ICP-AES) — Section 10
Procedure  (Con't)

10.1.4.3    Preparation Method/Code  (MSI)  (USEPA SW-846 Method 3051)

10.1.4.3.1     Add a representative  0.50 g  (ą0.01 g) of sample to the PTFE PFA
               vessel.

10.1.4.3.2     Add 10 mL of  concentrated nitric acid.  If a vigorous reaction
               occurs,  allow the  reaction  to  stop before capping  the vessel.

10.1.4.3.3     Cap the vessel, then  tighten using constant torque to 12
               ft/lbs,  according  to  the  manufacturer's direction.

10.1.4.3.4     Connect  the sample vessel to the overflow vessel using PTFE PFA
               tubing.

10.1.4.3.5     Weigh the vessel assembly to the nearest 0.01  g.

10.1.4.3.6     Place sample  vessels  in groups of 2  sample vessels or 6  sample
               vessels  in  the carousel,  evenly spaced around its  periphery in
                the microwave unit.   If  fewer  than  the recommended number of
                samples  are to be  digested  (i.e.,  3  samples  plus  1 blank) then
                the remaining vessels must  be  filled with  10 mL of nitric acid
                to achieve  the full  complement of  vessels.

10.1.4.3.7     Each  sample vessel must be  attached  to a clean, double-ported
                vessel  to collect  any sample expelled from the sample vessel  in
                the event of over  pressurization.   Assembly of the vessels into
                the carousel may be  done  inside or outside the microwave.
                Connect the overflow vessel to the center  well of  the oven.

10.1.4.3.8     The PB must have 0.5  mL of  reagent water and the same amount
                 (10 mL)  of acid that is  added to the samples.  The PB must
                later be diluted to  50 mL in the same manner as the samples.

10.1.4.3.9     Irradiate the 2 sample vessel  group  at 344  watts for 10
                minutes, or the 6-sample  vessel group at  574 watts for 10
                minutes.

10.1.4.3.10   This  program  brings  the samples to  175°C in  5.5 minutes;  the
                temperature remains  between 170-180°C for  the balance of the 10
                minute irradiation period.   The pressure  should peak at less
                than 6 atmospheres (atm)  for most samples.   The pressure may
                exceed these limits  in the  case of high concentrations of
                carbonate or organic compounds.  In these  cases,  the pressure
                will be limited by the relief pressure of  the vessel to 7.5
                 (ą0.7 atm).

10.1.4.3.11    Allow the vessels  to cool for  a minimum of  5 minutes before
                removing them from the microwave unit, with exhaust fan on.
                Allow the vessels to cool to room temperature before opening.
                The vessels must be carefully vented and uncapped in a fume
                hood.

10.1.4.3.12    Weigh each vessel  assembly.  If the  weight  of acid plus  the
                sample has decreased by more than 10% from the original weight,
                discard the digests.  Determine the reason for the loss.
                Losses typically are attributed to use of digestion time longer
                than ten minutes,  using too large of a sample, or having
                improper heating conditions.  Once the source of the losses has
                been corrected, prepare a new set of samples for digestion.

10.1.4.3.13    Sample  Filtration:  Shake  the sample  well to mix in any
                condensate within the digestion vessel before being opened.


 ILM05.2                       D-20/ICP-AES

-------
                                              Exhibit D (ICP-AES)  — Section 10
                                                              Procedure  (Con't)

               Filter the digestion vessel into a 50 mL glass volumetric flask
               through filter paper.  Rinse the sample digestion vessel, cap,
               connecting tube, and  (if venting occurred) the overflow vessel
               into the 50 mL glass flask.  Dilute to 50 mL.  The samples are
               now ready for analysis.  Concentrations so determined shall be
               reported as "total".

10.1.5   Non-Prepared Samples

10.1.5.1    Preparation Method/Code (NP1)

10.1.5.1.1     This code shall be used to report samples that are not digested
               prior to analysis  (e.g., dissolved metal samples that the
               Contractor was instructed not to digest).

10.1.5.1.2     This Preparation Method/Code shall also be used to report the
               non-prepared Method Detection Limit  (MDL).  The concentration
               of this MDL shall be used to determine the appropriate
               concentration qualifier for the results of non-prepared samples
               and instrument Quality Control  (QC) analyses.

10.2  Microwave Digestion Cleaning Procedure

10.2.1   Initial Cleaning of the PTFE PFA Digestion Vessels

10.2.1.1    Prior to first use - new vessels must be annealed before they are
            used.  A pretreatment/cleaning procedure must be followed.  This
            procedure calls for heating the vessels for 96 hours at 200°C.
            The vessels must be disassembled during annealing and the sealing
            surfaces  (the top of the vessel or its rim) must not be used to
            support the vessel during annealing.

10.2.1.2    Rinse in reagent water.

10.2.1.3    Immerse in 1:1 HC1 for a minimum of 3 hours after the cleaning
            bath has reached a temperature just below boiling.

10.2.1.4    Rinse in reagent water.

10.2.1.5    Immerse in 1:1 HN03 for a minimum of 3 hours after the cleaning
            bath has reached a temperature just below boiling.

10.2.1.6    The vessels are then rinsed with copious amounts of reagent water
            prior to use for any analyses under this contract.

10.2.2   Cleaning Procedure between Sample Digestions

10.2.2.1    Wash entire vessel in hot water using laboratory-grade non-phos-
            phate detergent.

10.2.2.2    Rinse with 1:1 nitric acid.

10.2.2.3    Rinse 3 times with reagent water.

10.3  Sample Analysis

10.3.1   Set up the instrument with proper operating parameters established in
         Section 9.1.  The instrument must be allowed to become thermally
         stable before beginning.  This usually requires at least 30 minutes
         of operation prior to calibration.

10.3.2   Initiate appropriate operating configuration of computer.


                                 D-21/ICP-AES                           ILM05.2

-------
Exhibit D (ICP-AES)  — Section 10
Procedure (Con't)

10.3.3   Profile and calibrate instrument according to instrument
         manufacturer's recommended procedures, using mixed calibration
         standard solutions such as those described in Section 7.2.4.5.1.

10.3.4   A minimum of two replicate exposures is required for standardization
         and all QC and sample analyses.  The average result of the multiple
         exposures for the standardization and all QC and sample analyses
         shall be used.
 ILM05.2                        D-22/ICP-AES

-------
                                             • Exhibit D (ICP-AES)  — Section 11
                                                 Data Analysis and Calculations

11.0  DATA ANALYSIS AND CALCULATIONS

11.1  Water/Aqueous Sample Calculation

      The concentrations determined in the digestate are to be reported in
      units of microgram per Liter  (pg/L):

      EQ. 2  Aqueous Sample Concentration
                                                  V
                        Concentration (pg/L) = C x —- x DF
       WHERE,      C    =   Instrument value in pg/L

                   VŁ   =   Final digestion volume  (mL)

                   Vt   =   Initial digestion volume (mL)

                   DF   =   Dilution Factor

11.2  Soil Sample Calculation

      The concentrations determined in the digestate are to be reported on the
      basis of the dry weight of the  sample, in units of milligrams per
      kilogram (mg/kg):

      EQ. 3  Soil Sample Concentration

                                                    C x V
                  Concentration  (dry wt.) (mg/kg) = 	 x DF
                                                    W x S
       WHERE,      C    =   Concentration (mg/L)

                   V    =   Final sample volume in Liters (L)

                   W    =   Wet sample weight  (kg)

                   S    =   % Solids/100 (see Exhibit D - Introduction to
                            Analytical Methods, Section 1.6).

                   DF   =   Dilution Factor

11.3  Adjusted Method Detection Limit  (MDL)/Adjusted Contract Required
      Quantitation Limit  (CRQL) Calculation

      To calculate the adjusted MDL or adjusted CRQL for water/aqueous
      samples, substitute the value of the MDL  (pg/L) or CRQL  (pg/L) into the
      "C" term in Equation 2 above.

      Calculate the adjusted MDL or adjusted CRQL for soil samples as follows:
                                 D-23/ICP-AES                           ILM05.2

-------
Exhibit D  (ICP-AES) — Section 11
Data Analysis and Calculations  (Con't)

      EQ.  4  Adjusted Soil MDL/Adjusted  Soil  CRQL  Concentration
                                                       W    V     i
         Adjusted Concentration (dry wt.) (mg/kg)  = Cx — x — x —  xDF
                                                       WR   VM    s


       WHERE,        C   =   MDL or CRQL concentration  (mg/kg)

                    WM   =   Minimum method required wet sample weight  (g)

                    WR   =   Reported wet sample weight  (g)

                    VM   =   Method required final sample volume  (mL)

                    VR   =   Reported final sample volume  (mL)

                    S   =   % Solids/100  (see Exhibit  D - Introduction to
                            Analytical Methods, Section 1.6).

                    DF  =   Sample Dilution Factor
 ILM05.2                        D-24/ICP-AES

-------
                                              Exhibit D (ICP-AES) — Section 12
                                                                 Quality  Control

12.0  QUALITY CONTROL  (QC)

12.1  Initial Calibration Verification  (ICV)

      The ICV standard shall be prepared in the same acid matrix as the
      calibration standards and in accordance with the instructions provided
      by the supplier.  If measurements exceed the control limits of 90%  (low)
      and 110%  (high), the analysis shall be terminated, the problem
      corrected, the instrument recalibrated, and the calibration reverified.
      Information regarding the ICV shall be reported on Form IIA-IN.

12.2  Continuing Calibration Verification (CCV)

      The CCV standard shall be prepared by combining compatible elements at a
      concentration equivalent to the mid-points of their respective
      calibration curves.  If the deviation of the CCV is greater than the
      control limits specified of 90%  (low) and 110%  (high), the analysis
      shall be  stopped, the problem corrected, the instrument recalibrated,
      the calibration verified, and the re-analysis of preceding 10 analytical
      samples or all analytical samples analyzed since the last  compliant
      calibration verification shall be performed for the analytes affected.
      Information regarding the CCV shall be reported on Form IIA-IN.

12.3  Contract  Required Quantitation Limit  (CRQL) Check Standard (CRI)

12.3.1   To verify linearity near the CRQL, a standard at the CRQL  (CRI) shall
         be prepared, -in  the same acid matrix as the calibration standards,
         and analyzed  at  the beginning and  end of each sample analysis run,
         immediately preceding the Interference Check Sample  (ICS) analyses,
         but not before the ICV.  In addition, the Contractor shall analyze
         the CRI at a  frequency of not less than once per 20 analytical
         samples1 per analysis run.  These analyses of the CRI sample shall be
         immediately followed by the ICS analyses.  [That is, the analytical
         run sequence  shall be CRI, ICS Solution A  (ICSA), ICS Solution  AB
          (ICSAB), CCV  and Continuing Calibration Blank  (CCB), in that order].

12.3.2   The CRI shall be run for every wavelength used for analysis, except
         those  for Al, Ba, Ca, Fe, Mg, Na,  and K.  Information regarding the
         CRI shall be  reported on Form IIB-IN.

12.3.3   If the percent recovery of the CRI falls outside the control limits
         of 70-130%  (50-150% for antimony,  lead, and thallium) for one or more
         analytes, the CRI shall be re-analyzed immediately for  those analytes
         only.  If the results of the re-analysis for those analytes fall
         within the control limits, no further corrective action is required.
         If the results of the re-analysis  for those analytes do not fall
         within the control limits, the analysis shall be terminated, the
         problem corrected, the instrument  recalibrated, the CRI analyzed, and
         the samples associated with the CRI re-analyzed.

12.4  Blank Analyses

      There are two different types of blanks required by this method.   The
      calibration blank is used in establishing the analytical curve while the
      Preparation Blank is used to monitor  for possible contamination.
      1As defined in Exhibit G,  CRI is an analytical sample.

                                 D-25/ICP-AES
ILM05.2

-------
Exhibit D (ICP-AES) — Section 12
Quality Control (Con't)

12.4.1   Initial and Continuing Calibration Blank  (ICB/CCB)

         The ICB and CCB are prepared with acids and reagent water.  If the
         absolute value of  the calibration blank  (ICB/CCB) result exceeds the
         CRQL  (see Exhibit  C), the analysis shall  be terminated, the problem
         corrected, the instrument recalibrated, the calibration verified, and
         re-analysis of the preceding 10 analytical samples or all analytical
         samples analyzed since the last compliant calibration blank shall be
         performed for the  elements affected.

12.4.2   Preparation Blank  (PB)

12.4.2.1    The PB shall contain all the reagents  and in the same volumes as
            used in processing the samples.  The PB shall  be carried through
            the complete procedure and  contain  the same acid concentration in
            the final  solution as the sample solution used for analysis.

12.4.2.2    At least one PB, consisting of reagent water processed through
            each sample preparation and analysis procedure (see Section 10),
            shall  be prepared and analyzed with every Sample Delivery  Group
             (SDG), or  with  each batch 2  of  samples  digested,  whichever  is  more
            frequent.

12.4.2.3    The first  batch of samples  in an SDG is to be  assigned to
            Preparation Blank one, the  second batch to  Preparation Blank  two,
            etc.  (see  Form  III-IN).  Each  Sample  Data Package  shall contain
            the results of  all PB analyses associated with the samples in that
            SDG.

12.4.2.4    The PB is  to be reported for each SDG  and used in all analyses to
            ascertain  whether  sample concentrations reflect  contamination in
            the following manner:

12.4.2.4.1     If  the  absolute value of the concentration  of the blank is less
                than  or equal  to  the  CRQL  (see Exhibit C),  no further action  is
                required.

12.4.2.4.2     If  any  analyte concentration in  the blank  is  above the  CRQL,
                the lowest concentration of that analyte in the associated
                samples shall  be  greater than  or equal  to  10  times  the  blank
                concentration.   Otherwise,  all  samples  associated  with  the
                blank,  with  the  analyte  concentration  less than 10  times  the
                blank concentration  and  above  the CRQL,  shall be  redigested and
                re-analyzed with  appropriate  new Quality Control  (QC)  for  that
                analyte.   The only exception  to  this  shall be an  identified
                field blank.  The sample concentration  is  not to be  corrected
                for the blank value.

12.4.2.4.3      If  the  concentration  of  the blank  is below the  negative CRQL,
                then  all samples  reported  below  10 times the  CRQL  associated
                with  the blank,  shall be redigested and  re-analyzed with
                appropriate new QC.

12.4.2.4.4      The values  for  the  PB shall be reported  on Form III-IN.
       2A  group  of  samples  prepared  at  the  same  time.

 ILM05.2                       D-26/ICP-AES

-------
                                              Exhibit D (ICP-AES) ~ Section 12
                                                        Quality  Control  (Con't)
12.5  Interference Check Sample (ICS)
12.5.1   The ICS is prepared by the analyst or obtained from USEPA, if
         available.

12.5.2   To verify interelement and background correction factors, the
         Contractor shall analyze and report the results for the  ICS, for all
         elements on the Target Analyte List  (TAL) and for all interferents
         (target and non-target), at the beginning and end of each analysis
         run, but not before the ICV.  In addition, the Contractor shall
         analyze and report the results for the ICS at a frequency of not less
         than once per 20 analytical samples3  per  analysis  run.   These
         analyses of the ICS shall be immediately followed by the analysis of
         a CCV/CCB pair.  The ICS solutions shall be obtained from USEPA, if
         available, and analyzed according to the instructions supplied with
         the ICS.  The Contractor shall not dilute the ICS more than is
         necessary to meet the linear range values of the instrument.

12.5.3   The ICS consists of two solutions: Solution A and Solution AB.
         Solution A consists of the interferents, and Solution AB consists of
         the analytes mixed with the interferents.  An ICS analysis consists
         of analyzing both solutions consecutively, starting with Solution A.

12.5.4   The analytical results of ICS Solution A  (ICSA)  shall fall within the
         control limit of ą2 times the CRQL of the analyte's true value or
         ą20% of the analyte's true value, whichever is greater  (the true
         value shall be zero unless otherwise stated) in the ICSA.  For
         example, if the analysis result (s) for Arsenic  (CRQL = 15 ug/L, ICSA
         true value = 0 ug/L) in the,ICSA analysis during the run is 29 ug/L,
         then the analytical result for Arsenic falls within the  ą2 times the
         CRQL window for Arsenic in the ICSA.  If the analytical  results of
         the ICSA do not fall within the control limits, the analysis shall be
         terminated, the problem corrected, the instrument recalibrated, and
         re-analysis of the analytical samples analyzed since the last
         compliant ICSA shall be performed.  For analytes with CRQLs less than
         5000 ug/L, the ICSA results shall be reported from an undiluted
         sample analysis.

12.5.5   Results for the ICS Solution AB (ICSAB) during the analytical runs
         shall fall within the control limit of ą2 times the CRQL of the true
         value or ą20% of the true value, whichever is greater, for the
         analytes included in the ICSAB.  If the analytical results of the
         ICSA do not fall within the control limits, the analysis shall be
         terminated, the problem corrected, the instrument recalibrated, and
         re-analysis of the analytical samples analyzed since the last
         compliant ICSAB shall be performed.

         NOTE: The control limits and concentrations for the ICSAB are being
         monitored.  These may be adjusted to provide greater control of
         interferences.

12.5.6   If true values for analytes contained in the ICS are not supplied
         with the solutions, the mean shall be determined by initially
         analyzing the ICS at least five times repetitively for the particular
         analytes.  This mean determination shall be made during  an analytical
         run where the results for the previously supplied ICS met all
         contract specifications.  Additionally, the results of this initial
         mean determination shall be used as the true value for the lifetime
         of that solution  (i.e., until the solution is exhausted).  Only if
      3As defined in Exhibit G,  ICSA and ICSAB are analytical samples.

                                 D-27/ICP-AES                            ILM05.2

-------
Exhibit D (ICP-AES) — Section 12
Quality Control (Con't)

         the' ICS solutions are not available from USEPA, independent Check
         Samples shall be prepared with interferent and analyte concentrations
         at the levels specified in Table 1 - Interferent and Analyte
         Elemental Concentrations Used for ICP-AES Interference Check Sample
          (ICS).  The mean value and standard deviation shall be established by
         initially analyzing  the Check Samples at least five times
         repetitively  for each parameter on Form IVA-IN.  Results shall fall
         within the  control limit of ą2 times the CRQL of the established mean
         value or  ą20% of the established mean value, whichever is greater.
         The mean  and  standard deviation shall be reported  in the raw data.
         Results from  the ICS analyses shall be reported on Form IVA-IN for
         all Inductively Coupled Plasma - Atomic Emission Spectroscopy  (ICP-
         AES)  analytes.

12.6  Spike Sample Analysis

12.6.1   The spike sample analysis is designed to provide information about
         the effect  of the  sample matrix on the digestion and/or measurement
         methodology.  If a digestion is performed, the spike is added before
         the digestion (i.e., prior to the addition of other reagents).  At
         least one spike sample analysis  (matrix spike) shall be performed on
         each  group  of samples of a similar matrix type  (i.e., water, soil) or
         for each  SDG.4

12.6.2   If the spike  analysis is performed on the same sample that is chosen
         for the duplicate  sample analysis, spike calculations shall be
         performed using the  results of the sample designated as the "original
          sample"  (see  Section 12.7).  The average of  the duplicate  results
          cannot be used  for the purpose of determining percent recovery.
          Samples identified as  field blanks and Performance Evaluation  (PE)
          samples shall not  be used  for spiked  sample  analysis.  USEPA may
          require that  a  specific  sample be used for the spike sample analysis.

12.6.3   The analyte spike  shall be added in the amount given in Table 2 -
          Spiking Levels  for Spike Sample Analysis, for  each element analyzed.

         NOTE: See Table  2  footnotes  for concentration  levels and
          applications.

12.6.4    If  the spike  recovery  is not at or within the  limits of 75-125%, the
          data  of all samples  received and associated  with  that spike sample
          shall be  flagged with the  letter "N"  on  Forms  IA/IB-IN  and VA-IN.  An
          exception to  this  rule  is  granted when the  sample  concentration
          exceeds the spike  added concentration by a  factor  of  four  or more.
          In  such an  event,  the data shall be  reported unflagged  even if the
          percent  recovery does  not  meet  the  75-125%  recovery  criteria.

12.6.5   When  the  matrix  spike  recovery falls  outside the control limits and
          the sample  result  does  not exceed  four times the  spike  added,  a post-
          digestion spike  shall be performed  for those elements that do  not
          meet  the  specified criteria  (exception:  Ag).  Note that  if a post-
          digestion spike analysis is  required for an  analyte,  the  same  EPA
          sample that was  used for the matrix  spike  analysis shall be used  for
          the post-digestion  spike analysis.   Spike  the  unspiked  aliquot of  the
          sample at two times the indigenous  level or  two  times  the  CRQL,
          whichever is greater.   Results  of  the post-digestion spike shall be
          reported on Form VB-IN.
       4USEPA may require additional  spike  sample  analyses,  upon  USEPA Regional
 CLP Project Officer (CLP PO)  request.

 ILM05.2                       D-28/ICP-AES

-------
                                              Exhibit D (ICP-AES)  — Section 12
                                                        Quality Control  (Con't)

12.6.6   In the instance where there is more than one spike sample per matrix
         per SDG, if one spike sample recovery is not within contract
         criteria, flag all the samples of the same matrix and method in the
         SDG.  Individual component percent recoveries are calculated as
         follows:

         EQ. 5  Spike Percent Recovery

                           % Recovery =  SSR " SR x 100
                                           SA


         WHERE,    SSR  =  Spiked  Sample Result

                   SR -  =  Sample  Result

                   SA   =  Spike Added

12.6.7   When sample concentration is less than the Method Detection Limit
          (MDL), use SR = 0 only for purposes of calculating percent recovery.
         The Spike Sample Results  (SSRs),  Sample Results  (SRs), Spike Added
          (SA), and percent recovery (positive or negative) shall be reported
         on Form VA-IN.

12.6.8   The units used for reporting SSRs will be identical to those used for
         reporting sample results on Form  IA-IN.

12.7  Duplicate Sample Analysis

12.7.1   One duplicate sample shall be  analyzed from each group of samples of
         a similar matrix type  (i.e., water, soil) or for each SDG. 5
         Duplicates cannot be averaged  for reporting on Form IA-IN.

12.7.2   Duplicate sample analyses are  required for percent solids.  Samples
         identified as field blanks and PE samples shall not be used for
         duplicate sample analysis.  USEPA may require that a specific sample
         be used for duplicate sample analysis.  The Relative Percent
         Difference  (RPD) for each component is calculated as follows:

         EQ. 6  Duplicate Sample Relative  Percent Difference


                              RPD =  I S  ~ D I  x 100
                                     (S+D)/2


          WHERE,    RPD =  Relative Percent Difference

                      S =  Sample  Result (original)

                      D =  Duplicate Result

12.7.3   The results of the duplicate sample analyses shall be reported on
         Form VI-IN.  A control limit of 20% for RPD shall be used for
         original and duplicate sample  values greater than or equal to five
         times the CRQL  (see Exhibit C).   A control limit of the CRQL value
         shall be entered in the "Control  Limit" column on Form VI-IN if
         either the sample or duplicate value is less than five times the
         CRQL.  If the sample and duplicate values are greater than or equal
      5USEPA may require additional duplicate sample analyses,  upon USEPA
Regional CLP PO request.

                                 D-29/ICP-AES                            ILM05.2

-------
Exhibit D (ICP-AES) — Section 12
Quality Control (Con't)

         to five times the CRQL, or if the sample and duplicate values are
         less than the CRQL,  the "Control Limit" field  is left empty.

12.7.4   If one result is above five times the CRQL level and the other is
         below, use the  CRQL  criteria to determine if the duplicate analysis
         is in control.   If both sample and duplicate values are less than the
         MDL, the RPD  is not  calculated on Form VI-IN.  For solid sample or
         solid duplicate results less than five times the CRQL, enter the
         value of the  CRQL, corrected for sample weight and percent solids,
          (i.e., original, not duplicate sample weight and percent solids), in
         the "Control  Limit"  column.  If the  duplicate  sample results are
         outside the control  limits, flag all the data  for samples received
         associated with that duplicate sample with an  "*" on Forms IA/IB-IN
         and VI-IN.  In  the instance where there is more than one duplicate
         sample per SDG,  if one duplicate result is not within contract
         criteria, flag  all samples of the same matrix  in the SDG.  The
         percent difference data will be used by USEPA  to evaluate the long-
         term precision  of the methods for each element.  Specific control
         limits for each element will be added to Form  VI-IN at a later date
         based on  these  precision  results.

12.8   Laboratory Control Sample  (LCS) Analysis

12.8.1   Water/aqueous and solid LCS shall be analyzed  for each analyte using
         the same  sample preparations, analytical methods, and Quality
         Assurance/Quality Control (QA/QC) procedures employed for the EPA
         samples received.

12.8.1.1    The aqueous  LCS solution  (LCSW) shall be obtained from USEPA  [if
            unavailable, the  ICV solution(s)  may be used].  One LCSW  shall be
            prepared  and analyzed  for every group of aqueous samples  in a  SDG,
            or  for each  batch of aqueous  samples  digested, whichever  is more
             frequent.

12.8.1.2    The USEPA  provided solid LCS  (LCSS) shall be prepared and analyzed
            using  each of the procedures  applied  to  the solid  samples received
             (exception:  percent solids determination not  required).   If  the
            USEPA  LCSS is unavailable, other  USEPA QC  Check  Samples or other
             certified materials may be used.   The  control  limits  for  these
            materials  and samples  must be  documented.   One LCSS  shall be
            prepared  and analyzed for every  group  of  solid  samples  in a  SDG,
             or  for each  batch of samples  digested, whichever is more  frequent.

12.8.2  All LCS and percent  recovery results shall be  reported on Form VII-
          IN.   If the  percent  recovery for  the LCSW  falls  outside  the  control
          limits of 80-120%  (exception: Ag  and Sb), the  analyses  shall be
          terminated,  the problem corrected,  and  the  samples  associated with
          that  LCSW redigested and re-analyzed with  appropriate new  QC.

12.8.3   If the results  for  the LCSS  fall  outside the  control limits
          established  by USEPA, the analyses  shall be  terminated,  the  problem
          corrected,  and the  samples associated with  that  LCSS  redigested and
          re-analyzed  with appropriate new QC.

12.9   ICP-AES  Serial  Dilution Analysis

12.9.1   Prior  to  reporting  concentration  data  for the  analyte elements,  the
          Contractor shall analyze and report  the  results  of  the  ICP-AES  serial
          dilution  analysis.   The ICP-AES  serial  dilution  analysis  shall  be
          performed on a sample from each  group of  samples  of a  similar matrix
          type  (i.e.,  water,  soil)  or  for  each SDG,  whichever is more  frequent.
 ILM05.2                       D-30/ICP-AES

-------
                                              Exhibit D  (ICP-AES) — Section 12
                                                         Quality  Control  (Con't)

         Samples identified as field blanks and PE samples shall not be used
         for serial dilution analysis.

12.9.2   If the analyte concentration is sufficiently high (minimally a factor
         of 50 above the MDL in the original sample), the serial dilution  (a
         five fold dilution) shall then agree within 10% of the  original
         determination after correction for dilution.  If the dilution
         analysis for one or more analytes is not within a control limit of
         10%, a chemical or physical interference effect must be suspected,
         and the data for all affected analytes in the samples received and
         associated with that serial dilution must be flagged with an "E" on
         Form VIII-IN and Forms IA/IB-IN.

12.9.3,,   The percent differences for each component are calculated as follows:

         EQ. 7  Serial Dilution Percent Differences

                                         I I - S I
                          % Difference = J-=-	^-1 x 100
          WHERE,     I  =  Initial Sample Result  (Instrument reading)

                     S  =  Serial Dilution Result  (Instrument reading x5)

12.9.4   In the instance where there is more than  one serial dilution per SDG,
         if one serial dilution result is not within contract criteria,  flag
         all the samples of the same matrix in the SDG.  Serial dilution
         results and  "E" flags shall be reported  on Form VIII-IN.

12.10 Method Detection Limit  (MDL) Determination

12.10.1  Before any field samples are analyzed under this contract,  the  MDLs
         shall be determined  for non-prepared analyses  (Preparation
         Method/Code  "NP1"),  each digestion procedure and instrument used,
         prior to the start of contract analyses,  and annually thereafter, and
         shall meet the levels specified in Exhibit C.

         An MDL study shall be performed after major instrument maintenance,
         or changes in instrumentation or instrumental  conditions  to verify
         the current  sensitivity of the analysis.

12.10.1.1   To determine the MDLs, the Contractor  shall run MDL studies
            following the procedures given in 40  CFR, Part  136.  The
            Contractor shall  prepare the MDL samples by each digestion
            procedure used and shall analyze these samples  on each instrument
            used.  The Contractor shall also analyze the non-prepared MDL
            samples on each instrument used.

12.10.1.2   The determined concentration of the MDL shall be less  than half
            the concentration of the CRQL listed  in Exhibit C.

12.10.1.3   The concentration of the non-prepared  MDL (Preparation Method/Code
            "NP1") shall be used to determine the  appropriate concentration
            qualifier for the results of non-prepared samples and  instrument
            QC analyses.

12.10.1.4   The results of the MDL determination studies shall be  forwarded to
            the USEPA Regional CLP PO, Sample Management Office  (SMO)", and
            Quality Assurance Technical Support  (QATS).

12.10.1.5   The MDL results shall be reported on Form IX-IN.


                                 D-31/ICP-AES                            ILM05.2

-------
Exhibit D (ICP-AES) — Section 12
Quality Control  (Con't)

12.11 Interelement Corrections

12.11.1  Before any field samples are analyzed under this contract, the
         interelement  correction factors shall be determined prior to the
         start of  contract  analyses and at least quarterly thereafter.
         Correction factors for spectral interference due to Al, Ca, Fe, and
         Mg shall  be determined for all ICP-AES instruments at all wavelengths
         used for  each analyte reported by ICP-AES.  Interelement correction
         factors shall also be reported for any other elements  (including
         those on  the  TAL)  that have been determined to interfere with the
         requested target analyte(s).

         NOTE: Depending on sample matrix and interferences, it may be
         necessary to  analyze interelement correction factors at a frequency
         greater than  quarterly and/or at multiple  concentrations comparable
         to the  sample interferent levels.

12.11.2  If the  instrument  was adjusted in any way  that may affect the ICP-AES
         interelement  correction  factors, the factors shall be  redetermined
         and the results submitted for use.  In addition, all data used  for
         the determination  of the interelement correction factors  shall  be
         available to  the USEPA during an on-site laboratory evaluation.
         Results from  interelement correction factors determination shall be
         reported  on Form XA-IN and Form XB-IN for  all  ICP-AES  analytes.

12.12 Linear Range Analysis Standard  (LRS)

12.12.1  A linear  range verification check standard shall be analyzed and ,
         reported  quarterly (i.e., January, April,  July and October)  for each
         element on  Form XI-IN.   The  standard shall be  analyzed during a
         routine analytical run performed under this contract.  The
         analytically  determined  concentration of this  standard shall be
         within  5% of  the  true value.  This concentration is the upper limit
         of  the  ICP-AES  linear range  beyond which results cannot be reported
         under this  contract without  dilution of the analytical sample.
 ILM05.2                       D-32/ICP-AES

-------
                                          Exhibit D (ICP-AES)  — Sections 13-16
                                                             Method Performance

13.0  METHOD PERFORMANCE

      Not applicable.

14.0  POLLUTION PREVENTION

      See Section 1.15 in Exhibit D - Introduction to Analytical Methods.

15.0  WASTE MANAGEMENT

      See Section 1.16 in Exhibit D - Introduction to Analytical Methods.

16.0  REFERENCES

16.1  US Environmental Protection Agency. Methods for Chemical Analysis of
      Water and Wastes. Method 200.7. December 1982.

16.2  US Environmental Protection Agency. Test Methods for Evaluating Solid
      Waste, Physical/Chemical Methods  (SW-846). Method 3050B. Third Edition,
      Update III. December 1996.

16.3  American Society for Testing and Materials. Standard Practice for Sample
      Digestion Using Closed Vessel Microwave Heating Technique for the
      Determination of Total Recoverable Metals in Water. D4309-91. October
      1991.

16.4  US Government Printing Office. 40 Code of Federal Regulations, Part 136,
      Section 1, Appendix B.

16.5  US Environmental Protection Agency.  Test Methods for Evaluating Solid
      Waste, Physical/Chemical Methods  (SW-846).  Method 3015.  Third Edition,
      Update II.  September 1994.

16.6  US Environmental Protection Agency. Test Methods for Evaluating Solid
      Waste, Physical/Chemical Methods  (SW-846).  Method 3051.  Third Edition,
      Update II.  September 1994.
                                 D-33/ICP-AES                           ILM05.2

-------
Exhibit D (ICP-AES) — Section 17
Tables/Diagrams/Flowcharts

17.0  TABLES/DIAGRAMS/FLOWCHARTS

      TABLE 1:  Interferent and Analyte Elemental Concentrations Used for
                ICP-AES Interference Check Sample  (ICS)
Analytes
Ag
As
Ba
Be
Cd
Co
Cr
Cu
Mn
Ni
Pb
Sb
Se
Tl
V
Zn
(mg/L) Interferents
0.2 Al
0.1 Ca
0.5 Fe
0.5 Mg
1.0
0.5
0.5
0.5
0.5
1.0
0.05
0.6
0.05
0.1
0.5
1.0
(mg/L)
250
250
100
250












       NOTE:  ICS  Solution A (ICSA)  contains  the  interferents  at  the  indicated
       concentrations.   The ICSA may be  analyzed at  twice  the concentration
       indicated  when interferences are  present  at higher  concentrations  in  the
       sample.  ICS Solution AB (ICSAB)  contains all of  the analytes and
       interferents listed above at the  indicated concentrations.
 ILM05.2                       D-34/ICP-AES

-------
                                              Exhibit D  (ICP-AES) — Section 17
                                             Tables/Diagrams/Flowcharts  (Con't)
      TABLE 2:   Spiking Levels for Spike Sample Analysis

Element
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Water
(ug/L)
2,000
100
40
2,000
50
50
*
200
500
250
1,000
20
Soil'1'
(mg/kg)
*
20
8
400
10
10
*
40
100
50
*
4
Element
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc


Water
(ug/L)
*
500
500
*
50
50
*
50
500
500


Soil'1'
(mg/kg)
*
100
100
*
10
10
*
10
100
100


      *No spike required.  NOTE: Elements without spike levels, and not
designated with an asterisk, shall be spiked at appropriate levels.

      JThe levels shown indicate concentrations in the spike sample when the
wet weight of 1 gram of sample is taken for analysis.  Adjustment shall be
made to maintain these spiking levels when the weight of sample taken deviates
by more than 10% of these values.  Appropriate adjustment shall be made for
microwave digestion procedures where 0.5 grams of sample or 50 mL  (45 mL of
sample plus 5 mL of acid) or 55 mL  (50 mL of sample plus 5 mL of acid) of
aqueous sample are required for analysis.
      EQ. 8  Spiking Level Adjustment

                        mg/kg = pg/L x
fi'nal volume  (L)
sample weight (g)
                                 D-35/ICP-AES
                                  ILM05.2

-------
THIS PAGE INTENTIONALLY LEFT BLANK

-------
     EXHIBIT D - PART B

     ANALYTICAL METHODS
             FOR
INDUCTIVELY COUPLED PLASMA -
      MASS  SPECTROMETRY
         D-1/ICP-MS                            ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                            D-2/ICP-MS

-------
                   Exhibit D - Analytical Methods for ICP-MS

                               Table of Contents


Section                                                                     Page

1.0   SCOPE AND APPLICATION 	   5

2.0   SUMMARY OF METHOD	5

3.0   DEFINITIONS  	   5

4.0   INTERFERENCES 	   6
      4.1   Isobaric Elemental Interferences   	  6
      4.2   Abundance Sensitivit y 	  6
      4.3   Isobaric Polyatomic Ion Interference s 	  6
      4.4   Physical Interferences   	  6
      4.5   Memory Interferences   	  7

5.0   SAFETY   	7

6.0   EQUIPMENT AND SUPPLIES   	  8
      6.1   Glassware/Labware  	   8
      6.2   Inductively Coupled Plasma Mass Spectrometer  (ICP-MS)  	   8

7.0   REAGENTS AND STANDARDS   	  9
      7.1   Reagents   	'	9
      7.2   Standards	9
      7.3   Blanks  	13

8.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE   	  14
      8.1   Sample Collection  and Preservation   	  14
      8.2   Procedures for  Sample Storage  	  14
      8.3   Procedure for Sample Digestate Storage   	  14
      8.4   Contract Required  Holding Time   	  14

9.0   CALIBRATION  AND STANDARDIZATION 	  15
      9.1   Instrument Operating Parameters 	  15
      9.2   Inductively Coupled Plasma - Mass Spectrometry  (ICP-MS)
             Instrument Calibration Procedure 	  15
      9.3   Initial.Calibration Verification  (ICV)   	  16
      9.4   Continuing Calibration Verification  (CCV )   	  16
      9.5   Initial and Continuing Calibration Blank (ICB/CCB)    	  17

10.0  PROCEDURE	17
      10.1  Sample Preparation	  17
      10.2  Sample Analysis	18

11.0  DATA ANALYSIS AND CALCULATIONS   	20
      11.1  Recommended Elemental Equations 	  20
      11.2  Data Value Corrections   	20
      11.3  Multiple Monitored Isotopes  	  20
      11.4  Direct Analysis	20
      11.5  Prepared Sample Analysis   	  20
      11.6  Adjusted Method Detection Limit (MDL)/Adjusted Contract
            Required Quantitation Limit  (CRQL) Calculation    	  21

12.0  QUALITY CONTROL  (QC)   	21
      12.1  Tune Standard	21
      12.2  Initial Calibration Verification  (ICV)   	  21
      12.3  Continuing Calibration Verification  (CCV )   	  21
      12.4  Contract Required  Quantitation Limit  (CRQL) Check Standard
             (CRI)	21
      12.5  Blank  Analyses   	22
      12.6  Interference Check Sample (ICS) 	  23

                                   D-3/ICP-MS                             ILM05.2

-------
                   Exhibit D - Analytical Methods for ICP-MS

                           Table of Contents  (Con't)
Section              ,                                                       Page

      12.7  Spike Sample Analysi s	24
      12.8  Duplicate Sample Analysi s 	 25
      12.9  Laboratory Control Sample  (LCS) Analysis    	 26
      12.10 ICP-MS Serial Dilution Analysis	'	26
      12.11 Internal Standards   	 26
      12.12 Method Detection Limit  (MDL) Determination    .-	27
      12.13 Linear Dynamic Range  (LDR)   	 27

13.0  METHOD PERFORMANCE   	 28

14.0  POLLUTION PREVENTION   	 28

15.0  WASTE MANAGEMENT   	 28

16.0  REFERENCES   	28

17.0  TABLES/DIAGRAMS/FLOWCHARTS   	 29
      Table 1.     Isobaric Molecular-Ion Interferences    	 29
      Table 2.     Mass Choices for  Elements  that Must Be  Monitored
                   During the Analytical Run	32
      Table 3.     Recommended Elemental  Expressions for Isobaric
                   Interferences 	 33
      Table 4,     Internal  Standards   	34
      Table 5.     Spiking Levels  for  Spike Sample  Analysis    	  34
 ILM05.2                           D-4/ICP-MS

-------
                                             Exhibit D (ICP-MS) — Sections 1-3
                                                          Scope and Application
1.0   SCOPE AND APPLICATION
      This method provides procedures for 'the use of Inductively Coupled
      Plasma - Mass Spectrometry  (ICP-MS) to determine the concentration of
      dissolved and total recoverable elements in water/aqueous samples taken
      from hazardous waste sites.  This method is applicable to all metals in
      the Target Analyte List  (TAL) for ICP-MS in Exhibit C.

2.0   SUMMARY OF METHOD

      This method describes the multi-element determination of trace elements
      by Inductively Coupled Plasma - Mass Spectrometry  (ICP-MS).  Sample
      material in solution is  introduced by nebulization into a radio
      frequency plasma where energy transfer processes cause desolvation,
      atomization, and ionization.  The ions are extracted from the plasma
      through a differentially pumped vacuum interface and separated on the
      basis of their mass-to-charge ratio.  The separated ions are detected
      and the ion information processed by a data handling system.
      Interferences related to the technique must be recognized and corrected.
      Such corrections may include compensation for isobaric elemental
      interferences and interferences from polyatomic ions derived from plasma
      gas, reagents, or sample matrix.  Instrumental drift, as well as
      suppressions or enhancements of instrument response, must be corrected
      for the use of internal  standards.

3.0   DEFINITIONS

      See Exhibit G for a complete list of definitions.
                                   D-5/ICP-MS
ILM05.2

-------
Exhibit D  (ICP-MS) — Section 4
Interferences

4.0   INTERFERENCES            '

      Several types of interferences may cause inaccuracies in the
      determination of trace elements by Inductively Coupled Plasma - Mass
      Spectrometry  (ICP-MS).  To prevent this, appropriate steps must be taken
      in all analyses to ensure that potential interferences are taken  into
      account.  Possible interferences  are  in Sections 4.1 through 4.5.

4.1   Isobaric Elemental Interferences

      Isobaric Elemental Interferences  are  caused by isotopes of different
      elements which  form  singly or doubly  charged  ions of the same nominal
      mass-to-charge  ratio, and which cannot be  resolved by the mass
      spectrometer.   All elements  determined by  this method have, at minimum,
      one  isotope free of  isobaric elemental interference.  Of the analytical
      isotopes recommended for use with this method, only selenium-82
       (krypton) has an isobaric elemental interference.  If alternative
      analytical  isotopes  having higher natural  abundances are selected, in
      order to achieve greater sensitivity, an isobaric interference may
      occur.  All data obtained under such  conditions must be corrected by
      measuring the signal from another isotope  of  the interfering element and
      subtracting the appropriate  signal ratio from the isotope of interest.
      A record of this correction  process should be included with the report
      of the data.  It should be noted  that such corrections will only  be as
      accurate as the accuracy of  the isotope ratio used in the elemental
      equation for data  calculations.   Relevant  isotope ratios should be
      established prior  to the application  of any corrections.

 4.2   Abundance Sensitivity

      Abundance Sensitivity is a property defining  the degree to which  the
      wings of a  mass peak contribute to adjacent masses.  The abundance
      sensitivity is  affected by ion energy and  mass filter operating
      pressure.   Wing overlap interferences may  result when a small  ion peak
      is being measured  adjacent to a large one. The potential for  these
      interferences should be recognized and  the spectrometer resolution
       should be adjusted to minimize.

 4.3   Isobaric Polyatomic  Ion Interferences

      These  are caused by ions  consisting of  more than one atom which have  the
       same nominal mass-to-charge  ratio as  the  isotope of  interest,  and which
       cannot  be resolved by the mass  spectrometer.  These  ions are  commonly
       formed in the plasma or  interface system  from support gases  or sample
       components.  Most  of the  common  interferences have been identified  and
       are  listed  in Table 1 -  Isobaric  Molecular-Ion  Interferences,  with  the
       target  analytes affected.   Such  interferences must be recognized, and
       when they  cannot be avoided  by  the  selection  of  alternative  analytical
       isotopes,  appropriate corrections must  be  made  to  the data.   Equations
       for  the  correction of data  should be  established at  the time  of  the
       analytical  run sequence,  since  the  polyatomic ion  interferences will  be
       highly dependent  on the  sample matrix and chosen instrument  conditions.

 4.4    Physical  Interferences

       These are  associated with the physical  processes which  govern the
       transport  of the  sample into the  plasma,  sample  conversion  processes  in
       the plasma, and the transmission  of ions  through the plasma-mass
       spectrometer interface.   These  interferences  may result  in  differences
       between instrument responses for  the  sample  and the  calibration
       standards.   Physical interferences may occur  in the  transfer of solution
       to the nebulizer (e.g.,  viscosity effects),  at  the point  of aerosol
       formation and transport to the plasma (e.g.,  surface tension),  or during


 ILM05.2                           D-6/ICP-MS

-------
                                           Exhibit D (ICP-MS)  — Sections 4 & 5
                                                                          Safety

      the excitation and ionization processes within the plasma itself.  High
     .levels of dissolved solids in the sample may contribute to deposits of
      material on the extraction and/or skimmer cones.  Deposits can reduce
      the effective diameter of the orifices and therefore ion transmission.
      Dissolved solid levels not exceeding 0.2% (w/v) have been recommended to
      reduce such effects.  Internal standardization may be effectively used
      to compensate for many physical interference effects.  Internal
      standards ideally should have similar analytical behavior to the
      elements being determined.

4.5   Memory Interferences

      Memory Interferences result when isotopes of elements in a previous
      sample contribute to the signals measured in a new sample.  Memory
      effects, or carryover, can result from sample deposition on the sampler
      and skimmer cones, as well as from the buildup of sample material in the
      plasma torch and spray chamber.  The site where these effects occur is
      dependent on the element and can be minimized by flushing the system
      with a rinse blank between samples (see Section 7.3.3).  The possibility
      of memory interferences should be recognized within an analytical run
      and suitable rinse times should be used to reduce them.  The rinse times
      necessary for a particular element should be estimated prior to
      analysis.  This may be achieved by aspirating a standard, containing the
      elements corresponding to ten times the upper end of the linear range
      for a normal sample analysis period,  followed by analysis of the rinse
      blank at designated intervals.  The length of time required to reduce
      analyte signals to within a factor of ten of the Method Detection Limit
      (MDL) should be noted.  Memory interferences may also be assessed within
      an analytical run by using a minimum of three replicate integrations for
      data acquisition.  If the integrated signal values drop consecutively,
      the analyst should be alerted to the possibility of a memory effect, and
      should examine the analyte concentration in the previous sample to
      identify if it was high.  If a memory interference is suspected, the
      sample should be re-analyzed after a long rinse period.

5.0   SAFETY

      See Section 1.14 in Exhibit D - Introduction to Analytical Methods.
                                   D-7/ICP-MS                           ILM05.2

-------
Exhibit D  (ICP.-MS) — Section 6
Equipment and Supplies

6.0   EQUIPMENT AND SUPPLIES

      Brand names, suppliers, and part  numbers are  for  illustrative  purposes
      only.  No endorsement is implied.   Equivalent performance may  be
      achieved using equipment and  supplies  other than  those  specified  here,
      however, a  demonstration of equivalent performance meeting  the
      requirements of this Statement  of Work (SOW)  is the  responsibility of
      the  Contractor.   The Contractor shall  document any use  of alternate
      equipment or supplies in the  Sample Delivery  Group  (SDG) Narrative.

6.1   Glassware/Labware

6.1.1    250 milliliter (mL) beaker or  other appropriate vessel

6.1.2    Watch glasses

6.1.3    Funnels

6.1.4    Graduated cylinders

6.1.5    Various  volumetric flasks  (Type A)

6.1.6    Thermometer that covers  range  of 0-200°C

6.1.7    Whatman  No. 42 filter paper  or equivalent

6.1.8    Hot plate, block digester, or  other heating source capable  of
         maintaining  92-95°C.

6.1.9    Balances - Analytical Balance, 300  gram (g) capacity,  and minimum
          ą0.1  milligram  (mg).

6.2    Inductively Coupled Plasma  Mass Spectrometer  (ICP-MS)  consisting  of:

       ="     An instrument capable of  scanning the mass  range 5-250 atomic mass
             unit (amu)  with a minimum resolution capability of 1 amu peak
             width at 5% peak height and either a conventional or extended
             dynamic range detector.

       :"     A radio-frequency generator compliant with Federal Communications
             Commission  (FCC)  regulations.

       ™     A high purity (99.99%)  argon gas supply.

       :xl     A variable speed peristaltic pump to deliver sample solution to
             the nebulizer.

       ™     A mass-flow  controller on the nebulizer gas supply is required.
 ILM05.2                            D-8/ICP-MS

-------
                                                Exhibit D (ICP-MS) -- Section 7
                                                         Reagents and Standards

7.0   REAGENTS AND STANDARDS

7.1   Reagents

      Reagents may contain elemental impurities that might affect the
      integrity of analytical data.  Owing to the high sensitivity of
      Inductively Coupled Plasma - Mass Spectrometry (ICP-MS), high-purity
      reagents should be used whenever possible.  All acids used must be of
      ultra high-purity grade.  Suitable acids are available from a number of
      manufacturers or may be prepared by sub-boiling distillation.  Nitric
      acid is preferred for ICP-MS in order to minimize polyatomic ion
      interferences.  Several polyatomic ion interferences result when
      hydrochloric acid (HCl) is used, however, it should be noted that HC1 is
      required to maintain stability in solutions containing antimony and
      silver.  When HCl is used, corrections for the chloride polyatomic ion
      interferences must be applied to all data.

7.1.1    Reagent Water - The purity of this water must be equivalent to ASTM
         Type II water  (ASTM D1193-77).  Use this preparation for all
         reagents, standards, and dilutions of solutions.

7.1.2    Nitric Acid - Concentrated (specific gravity 1.41).

7.1.3    Nitric acid (1+1) - Add 500 milliliters (mL) cone. HNO3  to 400 mL of
         reagent water and dilute to 1 Liter  (L).

7.1.4    Nitric acid (1+9) - Add 100 mL cone, nitric acid to 400 mL of reagent
         water and dilute to 1 L.

7.1.5    Hydrochloric acid - Concentrated (specific gravity 1.19).

7.1.6    Hydrochloric acid (1+1) - Add 500 mL cone. HCl to 400 mL of reagent
         water and dilute to 1 L.

7.1.7    Hydrochloric acid (HCl) (1+4) - Add 200 mL cone. HCl to 400 mL
         reagent water and dilute to 1 L.

7.1.8    Ammonium hydroxide - Concentrated  (specific gravity 0.902).

7.1.9    Tartaric acid -  (CASRN 87-69-4).

7.2   Standards

7.2.1    Introduction

         The Contractor must provide all standards to be used with this
         contract.  These standards may be used only after they have been
         certified according to the procedure in Exhibit E, Section 8.0.  The
         Contractor must be able to verify that the standards are certified.
         Manufacturer's certificates of analysis must be retained by the
         Contractor and presented upon request.

7.2.2    Stock Standard Solutions

7.2.2.1     Stock standard solutions may be purchased from a reputable
            commercial source or prepared from ultra high-purity grade
            chemicals or metals  (99.99-99.999% pure).  All salts should be
            dried for 1 hour at 105°C unless otherwise specified.  Stock
            solutions should be stored in Fluorinated Ethylene Propylene  (FEP)
            fluorocarbon bottles.  Note that some metals, particularly those
            which form surface oxides, require cleaning prior to being
            weighed.  This may be achieved by pickling the surface of the
            metal in acid.  An amount in excess of the desired weight should
                                   D-9/ICP-MS
ILM05.2

-------
Exhibit D (ICP-MS) — Section 7
Reagents and Standards  (Con't)

            be pickled  repeatedly,  rinsed with water,  dried  and  weighed  until
            the desired weight  is  achieved.

7.2.2.2     Aluminum solution,  stock  [1 mL = 1000 micrograms  (pg) Al]  -  Pickle
            aluminum metal  in warm (1+1) HC1 to  an  exact  weight  of  0.100 g.
            Dissolve in 10  mL cone. HC1 and 2 mL cone,  nitric acid,  heating  to
            effect solution.  Continue heating until  the  volume  is  reduced to
            4 mL.  Cool and add 4  mLs of reagent water.   Heat until volume is
            reduced to  2 mL.  Cool and dilute to 100  mL with reagent water.

7.2.2.3     Antimony solution,  stock  (1 mL = 1000 ug  Sb)  - Dissolve 0.100 g
            antimony powder in  2 mL (1+1) nitric acid and 0.5 mL cone. HC1,
            heating to  effect solution.  Cool, add  20 mL  reagent water and
            0.15  g tartaric acid.   Warm the solution  to dissolve the white
            precipitate.  Cool  and dilute to 100 mL with  reagent water.

7.2.2.4     Arsenic solution, stock (1 mL = 1000 ug As) - Dissolve  0.1320 g
            As203 in a mixture of 50 mL reagent water and 1 mL cone, ammonium
            hydroxide.   Heat gently to dissolve. Cool and acidify  solution
            with  2 mL cone, nitric acid.  Dilute to 100 mL with  reagent  water.

7.2.2.5     Barium solution,  stock (1 mL = 1000  ug  Ba)  -  Dissolve 0.1437 g
            BaC03 in a  solution mixture of 10 mL reagent  water and  2 mL  cone.
            nitric  acid.  Heat  and stir to effect solution and degassing.
            Dilute  to  100 mL with  reagent water.

7.2.2.6     Beryllium solution,  stock (1 mL = 1000  ug Be)  -  Dissolve 1.965 g
            BeSO4:l"4H2O (DO NOT DRY)  in 50 mL reagent water.   Add 1 mL cone.
            nitric  acid.  Dilute to 100 mL with  reagent water.

7.2.2.7     Bismuth solution, stock (1 mL = 1000 ug Bi) - Dissolve  0.1115 g
            Bi2O3 in 5 mL cone,  nitric acid.   Heat to effect  solution.  Cool
            and dilute  to 100 mL with reagent  water.

7.2.2.8     Cadmium solution, stock (1 mL = 1000 ug Cd) - Pickle cadmium metal
             in (1+9)  nitric acid to an exact weight of 0.100 g.   Dissolve in 5
            mL (1+1)  nitric acid,  heating to effect solution.  Cool and dilute
             to 100  mL with reagent water.

7.2.2.9     Chromium  solution,  stock  (1 mL = 1000 ug  Cr)  - Dissolve 0.1923 g
             CrO3  in a solution  mixture of 10 mL  reagent water and 1 mL cone.
             nitric  acid.  Dilute to 100 mL  with  reagent water.

7.2.2.10    Cobalt  solution, stock (1 mL  =  1000  ug  Co)  -  Pickle  cobalt metal
             in (1+9)  nitric acid to an exact weight of 0.100 g.   Dissolve in 5
             mL (1+1)  nitric acid,  heating to  effect solution.  Cool and dilute
             to 100  mL with reagent water.

7.2.2.11    Copper  solution, stock (1 mL  =  1000  pg  Cu)  -  Pickle  copper metal
             in (1+9)  nitric acid to an exact  weight of 0.100 g.   Dissolve in 5
             mL (1+1)  nitric acid,  heating to  effect solution.  Cool and dilute
             to 100  mL with reagent water.

7.2.2.12     Indium  solution, stock (1 mL  =  1000  pg  In)  -  Pickle  indium metal
             in (1+1)  nitric acid to an exact  weight of 0.100 g.   Dissolve in
             10 mL (1+1) nitric acid, heating to effect solution.  Cool and
             dilute to 100 mL with reagent water.

7.2.2.13     Lead solution, stock  (1 mL =  1000  pg Pb)  - Dissolve  0.1599 g PbNO3
             in 5 mL (1+1) nitric acid.  Dilute  to 100 mL with reagent water.
 ILM05.2                           D-10/ICP-MS

-------
                                                Exhibit D  (ICP-MS) — Section 7
                                                 Reagents and Standards  (Con't)

7.2.2.14    Magnesium solution, stock  (1 mL = 1000 pg Mg) - Dissolve 0.1658 g
            MgO in 10 mL  (1+1) nitric  acid, heating to effect solution.  Cool
            and dilute to 100 mL with  reagent water.

7.2.2.15    Manganese solution, stock  (1 mL = 1000 ug Mn) - Pickle manganese
            flake in  (1+9) nitric acid to an exact weight of 0.100 g.
            Dissolve in 5 mL  (1+1) nitric acid, heating to effect solution.
            Cool and dilute to 100 mL  with reagent water.

7.2.2.16    Nickel solution, stock (1  mL = 1000 ug Ni) - Dissolve 0.100 g
            nickel powder in 5 mL cone, nitric acid, heating to  effect
            solution.  Cool and dilute to 100 mL with reagent water.

7.2.2.17    Scandium solution, stock  (1 mL = 1000 ug Sc) - Dissolve 0.1534
            Sc2O3  in  5 mL  (1+1) nitric acid,  heating  to  effect  solution.   Cool
            and dilute to 100 mL with  reagent water.

7.2.2.18    Selenium solution, stock  (1 mL = 1000 ug Se) - Dissolve 0.1405 g
            SeO2 in 20 mL reagent water and dilute to 100 mL with reagent
            water.

7.2.2.19    Silver solution, stock (1  mL = 1000 ug Ag) - Dissolve 0.100 g
            silver metal in 5 mL  (1+1) nitric acid, heating to effect
            solution.  Cool and dilute to 100 mL with reagent water.  Protect
            from the light.

7.2.2.20    Terbium solution, stock  (1 mL = 1000 pg Tb) - Dissolve 0.1176 g
            Tb4O7  in  5 mL  cone, nitric acid,  heating  to  effect  solution.   Cool
            and dilute to 100 mL with  reagent water.

7.2.2.21    Thallium solution, stock  (1 mL = 1000 pg Tl) - Dissolve 0.1303 g
            T1N03 in a solution mixture of 10 mL reagent water and 1 mL cone.
            nitric acid.  Dilute to  100 mL with reagent water.

7.2.2.22    Vanadium solution, stock  (1 mL = 1000 pg V) - Pickle vanadium
            metal in  (1+9) nitric acid to an exact weight of 0.100 g.
            Dissolve in 5 mL  (1+1) nitric acid, heating to effect solution.
            Cool and dilute to 100 mL  with reagent water.

7.2.2.23    Yttrium solution, stock  (1 mL = 1000 pg Y) - Dissolve 0.1270 g
            Y2O3  in 5  mL  (1+1)  nitric acid, heating to effect solution.   Cool
            and dilute to 100 mL with  reagent water.

7.2.2.24    Zinc solution, stock  (1 mL = 1000 pg Zn)  - Pickle zinc metal in
             (1+9) nitric acid to an  exact weight of 0.100 g.  Dissolve in 5 mL
             (1+1) nitric acid, heating to effect solution.  Cool and dilute to
            100 mL with reagent water.

7.2.3    Secondary Dilution Standards

         Prepare mixed secondary dilution standard solutions by  diluting the
         appropriate volumes of stock  standards with acidified reagent water
         to obtain the final volume.   Originating stock standards should be
         checked for the presence of  impurities which might influence the
         accuracy of the standard.   Freshly prepared standards should be
         transferred to acid-cleaned,  not previously used, FEP fluorocarbon
         bottles for storage and monitored periodically for stability.  Mixed
         secondary dilution standard  solutions may be purchased.  The
         purchased standards shall meet the requirements in Section 7.2.1.
                                  D-11/ICP-MS                            ILM05.2

-------
Exhibit D (ICP-MS) — Section 7
Reagents and Standards  (Con't)

7.2.4    Working Standards

7.2.4.1     Mixed Calibration Standard Solutions

            Care must be taken  in  the preparation of mixed  calibration
            standards to ensure that the elements are compatible  and  stable.
            Fresh calibration standards should be prepared  from mixed standard
            solutions every  two weeks or as needed.  Dilute the mixed
            standards to levels appropriate to the operating range  of the
            instrument  using reagent water containing 1%  (v/v)  nitric acid.
            The element concentrations in the calibration standards should be
            sufficiently high to produce good measurement precision and to
            accurately  define the  slope of the response curve.  If  the  direct
            addition procedure  is  being used, add internal  standards.

7.2.4.2     Internal Standard Solution

            Prepare mixed  standard by diluting 10 mL each of the  chosen
            element's stock  standards to 100 mL  with reagent water.  Use this
            solution  for additions to blanks, calibration standards,  and
            samples,  or dilute  by  an appropriate amount using 1%  (v/v)  nitric
            acid  if  the internal standards are being added  by a peristaltic
            pump.

7.2.4.3     Tuning Solution

            This  solution  is used  for instrument tuning and mass  calibration
            prior to  analysis.   Prepare mixed standard by diluting  beryllium,
            magnesium,  cobalt,  indium, and lead  stock standards to  100  pg/L
            with  1%  (v/v)  nitric acid.  Do not add  internal standard  to this
            solution.

7.2.4.4     Interference Check  Sample  (ICS)

            The  ICS  consists of two solutions: Solution A   (ICSA)  and  Solution
            AB (ICSAB).  ICSA consists of the interferents  and ICSAB  consists
            of the analytes  mixed  with the interferents.   If the  direct
            addition procedure  is  being used, add  internal  standards.

7.2.4.4.1       Solution A  -  Contains 100 milligrams per Liter (mg/L)  of
                aluminum,  calcium,  iron, magnesium,  potassium, sodium,
                phosphorus  (as orthophosphate),  sulfur  (as  sulfate), 200 mg/L
                carbon,  1000 mg/L chloride, and  2 mg/L molybdenum and titanium.

7.2.4.4.2       Solution AB - Contains all of the elements in Solution A plus
                all target analytes at  a  concentration  of  20 ug/L.

7.2.4.5     Contract Required Quantitation Limit (CRQL) Check Standard  (CRI)

             The concentrations  of  the  analytes  in  the CRI  shall be  at the
             CRQL.  Information  regarding  the  CRI shall  be  reported  on Form
             IIB-IN.

7.2.4.6     Method Detection Limit (MDL) Solution

             The MDL solution shall be  at  a  concentration  of 3 to 5  times the
             expected MDL.
 ILM05.2                           D-12/ICP-MS

-------
                                                Exhibit D (ICP-MS) — Section 7
                                                 Reagents and Standards  (Con't)
7.3   Blanks
      Three types of blanks are required for this method.  A calibration blank
      is used to establish the analytical calibration curve, the Preparation
      Blank  (PB)  (see Section 12.5.2) is used to assess possible contamination
      from the sample preparation procedure and to assess spectral background,
      and the rinse blank is used to flush the instrument between samples in
      order to reduce memory interferences.

7.3.1    Calibration Blank - Consists of 1% (v/v) nitric acid in reagent
         water.  If the direct addition procedure is being used, add internal
         standards.

7.3.2    Preparation Blank - Must contain all the reagents in the same volumes
         as used in preparing the samples.  The PB must be carried through the
         complete procedure and contain the same acid concentration in the
         final solution as the sample solution used for analysis.

7.3.3    Rinse Blank - Consists of 2% (v/v) nitric acid in reagent water.
                                   D-13/ICP-MS                            ILM05.2

-------
Exhibit D (ICP-MS) — Section 8
Sample Collection, Preservation, and Storage

8.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE

8.1   Sample Collection and Preservation

      All samples must be collected in glass or polyethylene containers.
      Water/aqueous samples must be preserved with nitric acid to pH less than
      2 immediately after collection.  All samples must be iced or
      refrigerated at 4°C  (ą2°C) from the time of collection until digestion.

8.1.1    Dissolved Metals

         For the  determination  of dissolved metals, the sample must be
         filtered through a 0.45 micrometer  ( ^n) pore diameter membrane
         filter at the  time of  collection or as soon as possible.  Use a
         portion  of the sample  to rinse the filter flask, discard this
         portion, and collect the required volume of filtrate.  Preserve the
         filtrate with  nitric acid to pH less than 2 immediately after
         filtration.

8.2   Procedures  for Sample Storage

      The samples must  be protected from light and refrigerated at 4°C  (ą2°C)
      from  the time of  receipt  until  60 days after the delivery of a complete,
      reconciled  data package to USEPA.  After 60 days the samples may be
      disposed of in a  manner that complies with all applicable regulations.

8.3   Procedure for Sample Digestate  Storage

      Sample digestates must be stored until 365 days after delivery of a
      complete, reconciled data package to USEPA.

8.4   Contract Required Holding Time

      The maximum holding  time  for metals is 180 days from Validated Time of
      Sample Receipt  (VTSR).
 ILM05.2                           D-14/ICP-MS

-------
                                                Exhibit D (ICP-MS) — Section 9
                                                Calibration and Standardization

9.0   CALIBRATION AND STANDARDIZATION

9.1   Instrument Operating Parameters

      Because of the differences between various makes and models of
      satisfactory instruments, no detailed operating instructions can be
      provided.  Instead, the analyst should follow the instructions provided
      by the manufacturer of the particular instrument.  The Method Detection
      Limit (MDL), precision, linear dynamic range, and interference effects
      must be investigated and established for each individual element on that
      particular instrument.  All measurements must be within the operational
      range of the instrument where corrections are valid.  It is the
      responsibility of the analyst to verify that the instrument
      configuration and operating conditions used satisfy the analytical
      requirements and to maintain Quality Control (QC) data confirming
      instrument performance and analytical results.

9.2   Inductively Coupled Plasma - Mass Spectrometry  (ICP-MS)  Instrument
      Calibration Procedure

9.2.1    Precalibration routine - The following precalibration routine must be
         completed prior to calibrating the instrument.

         Set up the instrument with proper operating parameters established in
         Section 9.1.  The instrument must be allowed to become stable prior
         to calibration.  Conduct any necessary mass calibration and
         resolution routines to bring peak width within 0.75 atomic mass unit
          (amu) at 5% peak height and adjust mass calibration to within 0.1 amu
         over the range of 6 to 210 amu.

         Demonstrate instrument stability and precision by analyzing the
         tuning solution a minimum of five times consecutively.  The percent
         relative standard deviation of the absolute signals for all analytes
         in the tuning solution must be less than 5%.

9.2.2    Internal Standardization

         Internal standardization must be used in all analyses to correct the
         instrument drift and physical interferences.  A list of acceptable
         internal standards is provided in Table 4 - Internal Standards.  For
         full range mass scans, a minimum of three internal standards shall be
         used.  The masses of the internal standards shall bracket the masses
         of the analyte.  Internal standards shall be present in all samples,
         standards, and blanks at identical levels.  This may be achieved by
         directly adding an aliquot of the internal standards solution to each
         sample, standard, and blank, or by mixing with the sample solution
         prior to nebulization using a second channel of the peristaltic pump
         and mixing coil.  The concentration of the internal standard should
         be sufficiently high for good precision and to minimize the
         possibility of correction errors if the internal standard is
         naturally present in the sample.  Depending on the sensitivity of the
         instrument, a concentration range of 20 ug/L to 200 ug/L of each
         internal standard is recommended.  Internal standards should be added
         to samples, standards, and blanks in a similar manner, in order for
         dilution effects to be disregarded.

9.2.3    Calibration

         Instruments shall be calibrated daily, once every 24 hours, or each
         time the instrument is set up.  The instrument standardization date
         and time shall be included in the raw data.   Calibration standards
         shall be prepared as in Section 7.2.4.1.  Calibrate the instrument
         with at least two standards, one of which must be a blank standard.
                                  D-15/ICP-MS
ILM05.2

-------
Exhibit D (ICP-MS) — Section 9
Calibration and Standardization  (Con't)

         A minimum of three  replicate  integrations are required  for data
         acquisition.  Use the average of the integrations  for instrument
         calibration and data reporting.

         NOTE: Any changes or corrections to the analytical system shall be
         followed by recalibration.

9.3   Initial Calibration Verification  (ICV)

9.3.1    Immediately after each  instrument has been calibrated,  the accuracy
         of the initial calibration  shall be verified and documented  for every
         analyte  by the analysis of  the  ICV solution (s)  for each mass used to
         report final results.

9.3.2    Only if  the ICV solution(s) is (are) not available  from  USEPA,  or
         where a  certified  solution  of an analyte  is not available from any
         source,  analyses shall  be conducted on an independent standard at a
         concentration other than that used for instrument  calibration, but  .
         within the calibration  range.   An independent standard  is defined as
         a  standard composed of  the  analytes from  a different source  other
         than those used  in  the  standards for instrument calibration.

9.3.3    The ICV  solution (s)  shall be  run at each  mass used for  reporting
         final results.   The values  for the ICV shall be reported on  Form IIA-
         IN.

9.4   Continuing  Calibration Verification  (CCV)

9.4.1    To ensure calibration" accuracy during each analysis run, one of the
         following standards shall be  used  for the CCV  for  each  mass  used for
         reporting final  results for each element, at a  frequency of  10% or
         every 2  hours  during an analysis run, whichever is more frequent.
         The  standard  shall also be  analyzed and  reported for each mass used
          for  reporting  final results for each element at the beginning of the
         run  and  after  the  last  analytical  sample. The  analyte  concentrations
          in the CCV  standard(s)  shall  be different from  the concentrations  for
          the  ICV  and  shall  be one of the following solutions at  or near one-
         half  of  the  calibration standard:

          <*:    USEPA Solutions

          Ť=    NIST Standards

          11    A Contractor-prepared standard solution

          The same CCV standard shall be used  throughout  the analysis  runs  for
          a Sample Delivery Group (SDG) of samples received.

 9.4.2    Each CCV analyzed shall reflect the conditions  of  analysis  of all
          associated analytical samples  (the preceding 10 analytical  samples  or
          the preceding analytical samples up  to  the  previous CCV).   The
          duration of analysis,  rinses, and other related operations  which may
          affect the  CCV measured result may not  be applied to the CCV to  a
          greater extent than the extent applied to the  associated analytical
          samples.  For instance, the difference  in time between a CCV analysis
          and the blank immediately following  it,  as  well as the difference  in
          time between the CCV and the analytical sample immediately preceding
          it,  may not exceed the lowest difference in time between any two
          consecutive analytical samples associated with the CCV.

 9.4.3     Information regarding the CCV shall be  reported on Form IIA-IN.
 ILM05.2                           D-16/ICP-MS

-------
                                          Exhibit D (ICP-MS)  — Sections 9 & 10
                                                                       Procedure

9.5   Initial and Continuing Calibration Blank (ICB/CCB)

      A calibration blank shall be analyzed for each mass used for reporting
      final results for each element immediately after every ICV and CCV, at a
      frequency of 10% or every 2 hours during the run, whichever is more
      frequent.  The blank shall be analyzed at the beginning of the run and
      after the last analytical sample.

      NOTE: A CCB shall be analyzed immediately after the last CCV, and the
      last CCV shall be analyzed immediately after the last analytical sample
      of the run.  The results of the calibration blanks shall be reported on
      Form III-IN.

10.0  PROCEDURE

10.1  Sample Preparation

10.1.1   If insufficient sample amount  (less than 90% of the required amount)
         is received to perform the analyses, the Contractor shall contact the
         Sample Management Office (SMO) to inform them of the problem.  SMO
         will contact the Region for instructions.  The Region will either
         require that no sample analysis be performed or will require that a
         reduced volume be used for the sample analysis.  No other changes in
         the analysis will be permitted.  The Contractor shall document the
         Region's decision in the Sample Delivery Group  (SDG) Narrative.

10.1.2   If multiphase samples (e.g., two-phase liquid sample, oily
         sludge/sandy soil sample) are  received by the Contractor, the
         Contractor shall contact SMO to apprise them of the type of sample
         received.  SMO will contact the Region.  If all phases of the sample
         are amenable to analysis, the  Region may require the Contractor to do
         any of the following:

          ™:    Mix the sample and analyze an aliquot from the homogenized
               sample.

          ™    Separate the phases of the sample and analyze one or more of
               the phases, separately.  SMO will provide EPA sample numbers
               for the additional phases, if required.

          ™    Do not analyze the sample.

10.1.2.1    If all of the phases are not amenable to analysis  (i.e., outside
            scope), the Region may require the Contractor to do any of the
            following:

            ^     Separate the phases and analyze the phase(s) that is (are)
                  amenable to analysis.  SMO will provide EPA  sample numbers
                  for the additional phases, if required.

            :J[:     Do not analyze the sample.

10.1.2.2    No other changes in the analyses will be permitted.  The
            Contractor shall document the Region's decision in the SDG
            Narrative.

10.1.3   Sample Preparation Procedures

10.1.3.1    Direct Analyses - Preparation Method/Code  (NP1)

10.1.3.1.1     For the analysis of dissolved analytes in water samples,
               transfer 20 milliliters  (mL) of the filtered, acid-preserved
               sample to a clean, closeable container such as a centrifuge


                                  D-17/ICP-MS                           ILM05.2

-------
Exhibit D (ICP-MS) — Section 10
Procedure (Con't)

                tube.  Add sufficient  (1+1) nitric acid to make  the  sample  1%
                (v/v)  acid.   If  the  direct addition procedure  is being used,
                add internal  standards, close the container  and  mix.  The
                sample is  now ready  for analysis.  If  a precipitate  is formed
                during acidification,  transport, or storage, sample  digestion
                is required as described  in Section 10.1.3.2.

10.1.3.1.2      This Preparation Method/Code shall also be used  to report the
                direct analysis  Method Detection Limit (MDL).  The
                concentration of this  MDL shall be used to determine the
                appropriate concentration qualifier for the  results  of non-
                prepared samples and instrument Quality Control  (QC)  analyses.

10.1.3.1.3      Prior  to the  analysis  of  aqueous samples for total recoverable
                analytes (excluding  CLP Quarterly Blind (QB) samples),
                determine  the turbidity and report the results of  these
                measurements, in Nephelolometric Turbidity Units (NTU),  in  the
                raw data.   If the turbidity is  less than 1 NTU,  direct analysis
                of the sample shall  be performed using the procedure in  Section
                10.1.3.1.1.  If  the  turbidity is greater than  or equal to 1
                NTU,  digest the  sample as follows:

10.1.3.2     Preparation Method/Code (HW2)

             Shake and transfer  a 100  mL  aliquot of the sample to  a  250  mL
             heating  vessel,  add 2 mL  (1+1) nitric acid and  1  mL of  (1+1)
             hydrochloric acid (HC1) to the sample.  Cover with  a  ribbed watch'
             glass and heat on either  a hot plate, block digester, or
             equivalent  heating  source which is adjustable and capable of
             maintaining a temperature of 92-95°C  for  2 hours, or  until  the
             sample volume is reduced  to  about  20 mL  (DO NOT BOIL).   Cover  with
             a watch  glass to prevent  additional evaporation and reflux  for 30
             minutes.   Cool sample,  transfer to a  50 mL volumetric flask, and
             adjust  sample volume to 50 mL with reagent water.  Mix  and  allow
             any solids  present  to settle by gravity overnight or  centrifuge
             (if after settling  or centrifuging, the sample  contains suspended
             solids,  a portion of the  sample may be  filtered prior to
             analysis).

10.1.3.2.1      Prior to analysis, adjust the chloride concentration by
                pipetting 20 mL of the digestate  into  a 50  mL  volumetric flask
                and dilute to volume with reagent  water and mix.  If the direct
                addition method is being  used,  add internal  standards and mix.
                The sample is now ready for  analysis.

10.2   Sample Analysis

10.2.1   For every  new or unusual matrix, it is highly recommended  that a
          semi-quantitative analysis be carried out  to screen  for  high element
          concentrations.   Information gained  from this may  be used  to prevent
          potential  damage to the detector during  sample analysis  and to
          identify elements which may be  higher than the linear  range.   Matrix
          screening may be carried out by diluting the sample  by a factor of
          500 and analyzing in semi-quantitative mode.  The  sample should also
          be screened for background levels  of  all  elements  chosen for use  as
          internal standards in order to  prevent bias  in the calculation of
          analytical data.

 10.2.2   Initiate instrument operating configuration.  Tune and calibrate  the
          instrument for the analytes of  interest.   Establish  instrument
          software run procedures for quantitative analysis.  For  all sample
          analyses,  a minimum of three replicate  integrations  are  required  for
 ILM05.2                           D-18/ICP-MS

-------
                                               Exhibit D (ICP-MS) — Section 10
                                                              Procedure  (Con't)

         data acquisition.  Use the average of the integrations  for data
         reporting.

10.2.3   The rinse blank should be used to flush the system between samples.
         Allow sufficient time to remove traces of the previous  sample or a
         minimum of one minute.  Samples should be aspirated for a sufficient
         period of time to obtain a stable response prior to the collection of
         data.

10.2.4   Samples having concentrations higher than the established linear
         dynamic range should be diluted into range and re-analyzed.  The
         sample should first be analyzed for the trace elements, protecting
         the detector from the high concentration elements, if necessary, by
         the selection of appropriate scanning windows.  The sample should
         then be diluted for the determination of the remaining  elements.
         Alternatively, the dynamic range may be adjusted by selecting an
         alternative isotope of lower natural abundance, provided QC data for
         that isotope have been established.  The dynamic range must not be
         adjusted by altering instrument conditions to an uncharacterized
         state.

10.2.5   All masses which might affect data quality must be monitored during
         the analytical run.  At a minimum, those masses prescribed in Table 2
         - Mass Choices for Elements that Must Be Monitored During the
         Analytical Run, must be monitored in the same scan that is used for
         the collection of the data.  This information should be used to
         correct the data for identified interferences.

10.2.6   During the analysis of samples, the laboratory must comply with the
         required QC described in Section 12.  For the determination of
         dissolved analytes or the direct analysis of aqueous samples with
         turbidity less than 1 NTU, the Preparation Blank  (PB) and Laboratory
         Control Sample  (LCS) are not required.
                                  D-19/ICP-MS                           ILM05.2

-------
Exhibit D  (ICP-MS) — Section 11
Data Analysis and Calculations

11.0  DATA ANALYSIS AND CALCULATIONS

11.1  Recommended Elemental Equations

      Elemental expressions recommended  for sample data calculations are
      listed in Table 3 - Recommended Elemental Expressions  for  Isobaric
      Interferences.  Do not report element concentrations below the
      determined Method Detection Limit  (MDL).

11.2  Data Value Corrections

      'Data values should be corrected for  instrument drift or  sample matrix
      induced interferences by  the application of internal standardization.
      Corrections for characterized spectral  interferences should be applied
      to  the data.  Chloride interference  corrections  should be  made on all
      samples, regardless of the addition  of  hydrochloric acid (HC1),  as  the
      chloride ion  is a common  constituent of environmental  samples.

11.3  Multiple Monitored Isotopes

      If  an element has more than one monitored isotope, examination of the
      concentration calculated  for each  isotope or the isotope ratios  will
      provide useful  information in detecting a possible spectral
      interference.   Consideration should  therefore be given to both primary
      and secondary isotopes in the evaluation of  sample concentration.   In
      some cases, secondary  isotopes may be  less sensitive or more  prone  to
      interferences than the primary recommended isotopes, therefore
      differences between the results do not  neces_sarily indicate a problem
      with data  calculated  for  the primary isotopes.

11.4  Direct Analysis

      EQ. 1  Non-Prepared Sample Concentration


                           Concentration (ug/L)= C x DF
        WHERE,       C    =    Instrument value in ug/L.   (The average of all
                             replicate  integrations).

                    DF   =    Dilution Factor
 11.5  Prepared Sample Analysis

       EQ.  2  Prepared Sample Concentration
                                                V    V
                     Concentration (yg/L) = C x — x —- x DF
       WHERE,       C    =   Instrument value in pg/L  (The  average of all
                             replicate integrations).

                    Vf   =   Final digestion volume (50 mL)

                    Y!   =   Initial digestion  volume  (100  mL)

                    DF   =   Dilution Factor


 ILM05.2                           D-20/ICP-MS

-------
                                         Exhibit D (ICP-MS) — Sections 11 & 12
                                                                 Quality Control

11.6  Adjusted Method Detection Limit  (MDL)/Adjusted Contract Required
      Quantitation Limit  (CRQL) Calculation

      To calculate the adjusted CRQL or adjusted MDL, multiply the value of
      the CRQL  (pg/L) or MDL  (ug/L) by the sample dilution  factor.'

12.0  QUALITY CONTROL (QC)

12.1  Tune Standard

      The Tune Standard shall be prepared in the same acid  matrix as the
      calibration standards and analyzed at least 5 times consecutively.  If
      the peak width at 5% peak height is not within 0.75 atomic mass units
      (amu) for each isotope, the mass calibration is not within 0.1 amu over
      the range of 6 to 210 amu, or the percent Relative Standard Deviation
      (%RSD)of the absolute signals of the analytes exceeds 5%, the analysis
      shall be terminated, the problem corrected, and the instrument re-tuned.
      All sample results  reported must be associated with an instrument tune
      that meets these requirements.

12.2  Initial Calibration Verification (ICV)

      The ICV Standard shall be prepared in the same acid matrix as the
      calibration standards and in accordance with the  instructions provided
      by the supplier.  If measurements exceed the control  limits of 90%  (low)
      and 110%  (high), the analysis shall be terminated, the problem
      corrected, the instrument recalibrated, and the calibration reverified.
      Information regarding the ICV shall be reported on Form IIA-IN.

12.3  Continuing Calibration Verification  (CCV)

      The CCV standard shall be prepared by combining compatible elements at a
      concentration equivalent to the mid-points of their respective
      calibration curves.  If the deviation of the CCV  is greater than the
      specified control limits of 90%  (low) and 110%  (high), the analysis
      shall be stopped, the problem corrected, the instrument recalibrated,
      the calibration verified, and re-analysis of the  preceding 10 analytical
      samples or all analytical samples analyzed since  the  last compliant
      calibration verification shall be performed for the elements affected.
      Information regarding the CCV shall be reported on Form IIA-IN.

12.4  Contract Required Quantitation Limit  (CRQL) Check Standard (CRI)

12.4.1   To verify linearity near the CRQL, a standard  at the CRQL  (CRI) shall
         be prepared, in  the  same acid matrix as the calibration standards,
         and analyzed at  the beginning and end of each  sample analysis run,
         but not before the ICV.  In addition, the contractor shall analyze
         the CRI at a frequency of not less than once per 20 analytical
         samples1 per analysis run.  The initial analysis of the CRI shall be
         immediately followed by the Interference Check Samples  (ICS)
         analyses.

12.4.2   The CRI shall be run for every required isotope used for the analysis
         of all Inductively Coupled Plasma - Mass Spectrometry  (ICP-MS)
         analytes.  Information regarding the CRI shall be  reported on Form
         IIB-IN.

12.4.3   If the percent recovery of the CRI falls outside the control limits
         of 70-130%  (50-150%  for cobalt, manganese, and zinc) for one or more
         analytes, the CRI shall be re-analyzed immediately for those analytes
      1As defined in Exhibit G, CRI is an analytical sample.

                                   D-21/ICP-MS                            ILM05.2

-------
Exhibit D (ICP-MS) — Section 12
Quality Control  (Con't)

         only.   If the  results of the re-analysis  for those analytes fall
         within  the  control  limits, no further corrective action is required.
         If the  results of the re-analysis for those analytes do not fall
         within  the  control  limits, the analysis shall be terminated, the
         problem corrected,  the instrument recalibrated, the CRI analyzed, and
         the samples associated with the CRI re-analyzed.

12.5  Blank Analyses

      There are  two  different types of blanks required by this method.  The
      calibration blank is used in establishing the analytical curve while the
      preparation blank is used to monitor for possible  contamination.

12.5.1   Initial and Continuing Calibration Blank  (ICB/CCB)

         The ICB and CCB are prepared with acid and reagent water.  If  the
         absolute value of  the calibration blank  (ICB/CCB) result  exceeds the
         CRQL  (see Exhibit  C), the analysis  shall  be  terminated, the problem
         corrected,  the instrument recalibrated,  the  calibration verified, and
         re-analysis of the  preceding 10 analytical samples or all analytical
         samples analyzed since the last compliant calibration blank shall be
         performed  for the  elements affected.

12.5.2   Preparation Blank  (PB)

12.5.2.1    The  PB  shall contain  all the reagents  and in the  same  volumes as
            used in processing the  samples.   The  PB  shall  be  carried through
             the  complete procedure  and  contain  the same acid  concentration  in
             the  final solution as the  sample solution used for  analysis.

12.5.2.2    At least one PB, consisting of reagent water processed through
             each sample preparation  and analysis  procedure (see  Section 10),
             shall be prepared and analyzed with  every Sample  Delivery  Group
             (SDG),  or with each batch 2 of samples digested,  whichever is more
             frequent.

12.5.2.3    The  first batch of samples  in  an SDG  is to be  assigned to
             Preparation Blank one,  the  second batch to Preparation Blank two,
             etc. (see Form III-IN).   Each Sample Data Package shall contain
             the results of all PB analyses associated with the  samples in that
             SDG.

12.5.2.4    The  PB is to be reported  for  each SDG and used in all  analyses to
             ascertain whether sample  concentrations reflect contamination in
             the following manner:

12.5.2.4.1      If the absolute value  of the  concentration  of  the blank is less
                 than or  equal to the CRQL (see Exhibit C),  no further action is
                 required.

12.5.2.4.2      If the analyte concentration  in  the blank is  above  the  CRQL,
                 the  lowest concentration of that analyte in the associated
                 samples  shall be greater than or equal to 10 times the blank
                 concentration.  Otherwise,  all samples, associated with the
                 blank,  with  the analyte concentration less than 10 times the
                 blank concentration and above the CRQL, shall be redigested and
                 re-analyzed  with appropriate new Quality Control (QC)  for that
                 analyte.  The only exception to this shall be an identified
                 field blank.  The sample concentration is not to be corrected
                 for  the blank value.
       2A group of samples prepared at the  same  time.

 ILM05.2                           D-22/ICP-MS

-------
                                               Exhibit D  (ICP-MS) — Section 12
                                                        Quality Control  (Con't)


12.5.2.4.3     If the concentration of the blank is below the negative CRQL,
               then all samples reported below 10 times the CRQL associated
               with the blank, shall be redigested and re-analyzed with
               appropriate new QC.

12.5.2.4.4     The values for the PB shall be reported on Form III-IN.

12.6  Interference Check Sample (ICS)

12.6.1   The ICS is prepared by the analyst "or obtained from USEPA, if
         available.

12.6.2   To verify corrections for elemental and polyatomic isobaric
         interferences, the Contractor shall analyze and  report the results
         for the ICS for all elements on the Target Analyte List  (TAL) and for
         all interferents  (target and non-target), at the beginning of each
         analysis run, but not before the ICV.  This analysis of the ICS shall
         be immediately followed by analysis of a CCV/CCB pair.  The ICS
         solutions shall be obtained from USEPA, if available, and analyzed
         according to instructions supplied with the ICS.  The Contractor
         shall not dilute the ICS  (for the higher concentration elements) more
         than is necessary to meet the linear range values of the instrument.

12.6.3   The ICS consists of two solutions: Solution A and Solution AB.
         Solution A consists of the interferents, and Solution AB consists of
         the analytes mixed with the interferents.  An ICS analysis consists
         of analyzing both solutions consecutively, starting with Solution A.

12.6.4   The analytical results of ICS Solution A (ICSA)  shall fall within the
         control limit of ą3 times the CRQL of the analyte's true value or
         ą20% of the analyte's true value  (the true value shall be zero unless
         otherwise stated) in the ICSA, whichever is greater.  If not, the
         analysis shall be terminated, the problem corrected, the instrument
         recalibrated, and re-analysis of the analytical  samples analyzed
         since the last compliant ICSA shall be performed.  The ICSA results
         for these analytes shall be reported from an undiluted sample
         analysis.

12.6.5   Results for the ICS Solution AB (ICSAB) during the analytical runs
         shall fall within the control limit of ą3 times  the CRQL of the true
         value or ą20% of the true value, whichever is greater, for the
         analytes included in the ICSAB.  If not, the analysis shall be
         terminated, the problem corrected, the instrument recalibrated, and
         re-analysis of the analytical samples analyzed since the last
         compliant ICSAB shall be performed.

         NOTE: The control limits and concentrations for  the ICSAB are being
         monitored.  These may be adjusted to provide greater control of
         interferences.

12.6.6   If true values for analytes contained in the ICS are not supplied
         with the solutions, the mean shall be determined by initially
         analyzing the ICS at least five times repetitively for the particular
         analytes.  This mean determination shall be made during an analytical
         run where the results for a previously supplied  ICS met all contract
         specifications.  Additionally, the results of this initial mean
         determination shall be used as the true value for the lifetime of
         that solution  (i.e., until the solution is exhausted).  Only if the
         ICS solutions are not available from USEPA, independent Check Samples
         shall be prepared with interferent and analyte concentrations at the
         levels specified in Sections 7.2.4.4.1 and 7.2.4.4.2.  The mean value
         and standard deviation shall be established by initially analyzing


                                  D-23/ICP-MS                            ILM05.2

-------
Exhibit D (ICP-MS) — Section 12
Quality Control  (Con't)

         the Check Samples at least five times repetitively for each analyte
         listed  on Form  IVB-IN.  Results shall fall within the control limit
         of ą3 times the CRQL of the established mean value or ą20% of the
         established mean value, whichever is greater.  The mean and standard
         deviation shall be  reported in the raw data.  Results from the  ICS
         analyses shall  be reported on Form IVB-IN for all ICP-MS parameters.

12.7  Spike Sample Analysis

12.7.1   The spike sample analysis is designed to provide information about
         the effect of sample matrix on the digestion and/or measurement
         methodology.  If a  digestion is performed, the spike is added before
         the digestion  (i.e., prior to the addition of other reagents).   At
         least one spike sample analysis  (matrix spike) shall be performed for
         each SDG3.

12.7.2   If the  spike analysis is performed on the same sample that is chosen
         for the duplicate  sample analysis, spike calculations shall be
         performed using the results of the sample designated "original
         sample" (see  Section  12.8).  The average of the duplicate  results
         cannot  be used  for  the purpose of determining percent recovery.
         Samples identified as  field blanks and  Performance Evaluation  (PE)
         samples shall not  be  used  for spiked sample analysis.  USEPA may
         require that  a  specific  sample be used  for the spike sample analysis.

12.7.3   The analyte spike  shall be added in  the amount given in Table 5 -
         Spiking Levels  for Spike Sample Analysis,  for each element analyzed.

12.7.4   If the  spike  recovery  is not at or within the limits of 75-125%, the
         data  for all  samples  received and associated with that  spike  sample
         and  shall be  flagged with  the letter WN" on Forms IA/IB-IN and  VA-IN.
         An exception  to this  rule  is granted when  the sample concentration
         exceeds the  Spike  Added  (SA) concentration by a  factor  of  four  or
         more.   In such  an  event, the data shall be reported unflagged even  if
          the  percent  recovery does  not meet  the  75-125%  recovery criteria.

12.7.5   When  the matrix spike  recovery  falls outside the control limits and
          the  sample  result  does not exceed four  times the  spike  added, a post-
          digestion spike shall be performed  for  those elements  that do not
         meet the specified criteria.  Note  that if a post-digestion  spike
          analysis is  required for an  analyte, the  same EPA sample  that was
          used for the  matrix spike  shall be  used for  the  post-digestion  spike
          analysis.   Spike an unspiked aliquot of the  digestate  at  two  times
          the  indigenous  level or  two  times the CRQL,  whichever  is  greater.
          Results of the  post-digestion  spike  shall  be  reported  on  Form VB-IN.

12.7.6    In  the instance where there  is more  than  one spike sample  per matrix
          per SDG, if one spike sample recovery is  not within  contract
          criteria,  flag all the samples  in  the SDG.   Individual  component
          percent recoveries are calculated as follows:

          EQ.  3  Spike Percent Recovery


                            %Recovery =  SSR "  SR  x 100
                                    *      SA
       3USEPA may require additional spike  sample  analyses,  upon USEPA Regional
 CLP Project Officer (CLP PO) request.

 ILM05.2                           D-24/ICP-MS

-------
                                               Exhibit D  (ICP-MS) — Section 12
                                                        Quality  Control  (Con't)


         WHERE,  SSR = Spike Sample Result

                  SR = Sample Result

                  SA = Spike Added


12.7.7   When sample concentration is less than the Method Detection Limit
          (MDL), use SR = 0  only  for purposes of calculating percent recovery.
         The Spike Sample Results  (SSRs), Sample Results  (SRs),  Spike Added
          (SA), and percent  recovery  (positive or negative) shall be reported
         on Form VA-IN.

12.7.8   The units used for reporting SSRs will be identical to  those used for
         reporting sample results  on Form IA-IN.

12.8  Duplicate Sample Analysis

12.8.1   One duplicate sample shall be analyzed for each  SDG4.   Duplicates
         cannot be averaged for  reporting on Form IA-IN.

12.8.2   Samples identified as field blanks and PE samples shall not be used
         for duplicate sample analysis.   USEPA may require that  a specific
         sample be used for duplicate sample analysis.  The Relative Percent
         Difference  (RPD) for each analyte is calculated  as follows:

         EQ. 4  Duplicate Sample Relative Percent Difference

                              RPD  =  I S  " D I  x 100
                                     (S+D)/2
         WHERE,  RPD = Relative Percent Difference

                   S = Sample Result  (original)

                   D = Duplicate Result

12.8.3   The results of the duplicate sample analyses shall be reported on
         Form VI-IN.  A control limit of 20% for  RPD shall be used  for
         original and duplicate sample values  greater than or equal to five
         times the CRQL  (see Exhibit C).  A control limit equal  to  the CRQL
         shall be entered  in the "Control Limit"  column on Form  VI-IN if
         either the sample or duplicate value  is  less than five  times the
         CRQL.  If the sample and duplicate values are greater than or equal
         to five times the CRQL, or if the sample and duplicate  values are
         less than the CRQL, the "Control Limit"  field is left empty.

12.8.4   If one result is above five times the CRQL level and the other is
         below, use the CRQL criteria to determine if the duplicate analysis
         is in control.  If both sample and duplicate values are less than the
         MDL, the RPD is not calculated on Form VI-IN.  If the duplicate
         sample results are outside the control limits, flag all the data for
         samples received  associated with that duplicate sample  with an "*" on
         Forms IA/IB-IN and VI-IN.  In the instance where there  is  more than
         one duplicate sample per SDG, if one  duplicate result is not within
         contract criteria, flag all samples in the SDG.  The percent
      4USEPA may require additional duplicate sample analyses,  upon USEPA
Regional CLP PO request.

                                   D-25/ICP-MS                            ILM05.2

-------
Exhibit D  (ICP-MS) — Section 12
Quality Control  (Con't)

         difference  data will be used by USEPA  to evaluate the long-term
         precision of  the methods  for each  element.  Specific control limits
         for each element may be added  to Form  VI-IN at a later date based  on
         these precision results.

12.9  Laboratory  Control Sample  (LCS) Analysis

12.9.1   A water/aqueous LCS  (LCSW)  shall be analyzed for each analyte using
         the same sample preparations,  analytical methods, and Quality
         Assurance/Quality  Control (QA/QC)  procedures employed for  USEPA
         samples  received.

12.9.2   The LCSW solution  must be obtained from USEPA  (if unavailable, the
         ICV solution(s) may  be used).  One aqueous LCS shall be  prepared and
         analyzed for  each  group of samples in  an SDG, or for each  batch  of
         samples  digested,  whichever is more  frequent.

12.9.3   All LCSW and  percent recovery  results  shall be reported  on Form  VII-
         IN.   If  the percent  recovery for the LCSW  falls outside  the control
         limits  of  80-120%, the  analyses shall  be terminated, the problem
         corrected,  and the samples associated  with that LCSW redigested  and
         re-analyzed with appropriate new QC.

12.10 ICP-MS Serial  Dilution  Analysis

12.10.1  Prior to reporting concentration data  for the analyte elements,  the
         Contractor shall analyze  and report  the  results of  the  ICP-MS  serial
         dilution analysis.  The  ICP-MS serial  dilution analysis  shall  be
         performed  on  a sample  from each  SDG.   Samples  identified as  field
         blanks  and PE samples  shall not be used  for  serial  dilution  analysis.

12.10.2  If  the  analyte concentration is sufficiently high  (minimally a factor
         of  50 above the MDL  in the original  sample),  the  serial  dilution  (a
          five-fold  dilution)  shall then agree within  10% of  the  original
         determination after  correction for dilution.   If  the dilution
         analysis for  one or  more  analytes  is not within a  control  limit  of
          10%,  and the  internal  standards  in the original sample  met the
          contract criteria, an  interference effect  must be  suspected,  and the
          data for all  affected  analytes in  the  samples  received  and associated
          with that  serial dilution must be  flagged  with an  "E" on Forms IA/IB-
          IN  and VIII-IN.

 12.10.3  The percent differences  for each component are calculated  as  follows:

          EQ.  5  Serial Dilution Percent Difference
                                          I T _ O I
                           % Difference  = -L^	^-1 x 100
          WHERE,   I =   Initial  Sample Result  (Instrument Reading)

                   S =   Serial Dilution  Result (Instrument  Reading x5)

 12.10.4  In the instance where there is more than one serial dilution per SDG,
          if one serial dilution result is not within the contract criteria,
          flag all samples in the SDG.  Serial dilution results and "E" flags
          shall be reported on Form VIII-IN.

 12.11 Internal Standards

 12.11.1  The analyst shall monitor the responses from the internal standards
          throughout the sample  set being analyzed.  Ratios of the internal


 ILM05.2                           D-26/ICP-MS

-------
                                               Exhibit D (ICP-MS)  — Section 12
                                                        Quality Control (Con't)

         standard responses between isotopes should also be routinely
         monitored.  This information may be used to correct potential
         problems caused by mass dependent drift, errors incurred in adding
         the internal standards or increases in the concentrations of
         individual internal standards caused by background contributions from
         the sample.  The absolute response of any one internal standard must
         not deviate more than 60-125% of the original response in the
         calibration blank.  If deviations greater than these are observed,
         the laboratory shall monitor the calibration blank internal standard
         responses by re-analyzing the calibration blank.  If these are within
         the limits, the original sample shall be diluted by a factor of two,
         internal standards added, and the sample re-analyzed.  If the
         internal standard responses for the calibration blank are not within
         the limits, terminate the analysis, correct the problem,  recalibrate,
         verify the calibration and re-analyze all analytical samples analyzed
         since the last compliant calibration blank.  If the internal standard
         responses.for the diluted sample analysis are not within the limits,
         note this in the SDG Narrative.

12.12 Method Detection Limit (MDL) Determination

12.12.1  Before any field samples are analyzed under this contract,  the MDLs
         shall be determined for each instrument used, prior to the start of
         contract analyses, and annually thereafter, and shall meet the levels
         specified in Exhibit C.

         An MDL study shall be performed after major instrument maintenance,
         or changes in instrumentation or instrumental conditions to verify
         the current sensitivity of the analysis.

12.12.2  To determine the MDLs, the Contractor shall run MDL studies following
         the procedures given in 40 CFR, Part 136.  The Contractor shall
         prepare the MDL samples by each digestion procedure used and shall
         analyze these samples on each instrument used.  The Contractor shall
         also analyze non-prepared MDL samples on each instrument used.

12.12.3  The determined concentration of the MDL shall be less than half the
         concentration of the CRQL listed in Exhibit C.

12.12.4  The direct analysis MDL  (Preparation Method/Code "NP1")  shall be used
         to determine the appropriate concentration qualifier for the results
         of instrument QC.

12.12.5  The results of the MDL determination studies shall be forwarded to
         the USEPA Regional CLP PO, Sample Management Office  (SMO),  and
         Quality Assurance Technical Support  (QATS).

12.12.6  The MDL results shall be reported on Form IX-IN.

12.13 Linear Dynamic Range (LDR)

12.13.1  Before any field samples are analyzed under this contract,  the upper
         limit of the linear calibration range shall be established for each
         analyte by determining the signal responses from a minimum of three
         different concentration standards, one of which is close to the upper
         limit of the linear range.  The linear calibration range used for the
         analysis of samples shall be determined from the resulting data.   The
         upper LDR limit shall be an observed signal no more than 10% below
         the level extrapolated from lower standards.  Determined sample
         analyte concentrations that are greater than 90% of the determined
         upper LDR limit must be diluted and re-analyzed.  The LDRs must be
         verified whenever a change in instrument hardware operating
                                  D-27/ICP-MS                           ILM05.2

-------
Exhibit D  (ICP-MS) — Sections 13-16
Method Performance

         conditions indicate they should be redetermined, or verified
         quarterly.

13.0  METHOD PERFORMANCE

      Not  applicable.

14.0  POLLUTION PREVENTION

      See  Section  1.15  in Exhibit D - Introduction to Analytical Methods.

15.0  WASTE MANAGEMENT

      See  Section  1.16  in Exhibit D - Introduction to Analytical Methods.

16.0  REFERENCES

16.1  US Environmental  Protection Agency. Determination  of Trace Elements in
      Waters and  Wastes by  Inductively  Coupled  Plasma -  Mass Spectrometry.
      Method 200.8. Revision 5.4. 1994.
                                        (

16.2  US Environmental  Protection Agency. Test  Methods for Evaluating  Solid
      Waste, Physical/Chemical Methods  (SW-846).  Method  6020A.  Third Edition,
      Update IV-A.  1986.

16.3  US Government Printing Office.  40 Code  of Federal  Regulations, Part 136,
      Section  1,  Appendix B.
 ILM05.2                           D-28/ICP-MS

-------
                                               Exhibit D  (ICP-MS) —- Section \1
                                                     Tables/Diagrams/Flowcharts
17.0  TABLES/DIAGRAMS/FLOWCHARTS

Table 1.    Isobaric Molecular-Ion Interferences
Analyte
121Sb
123Sb
75As
138Ba
l37Ba
136Ba
l35Ba
134Ba
132Ba
l30Ba
9Be
u/1Cd
ll2Cd
lllCd
110Cd
113Cd
116Cd
106Cd
108Cd
52Cr
53Cr
5oCr
54Cr
59Co
63Cu
65Cu
208pb
206pb
Oxygen
PdO
AgO
CoO
SnO
SbO
SnO
SnO
SnO
SnO,
CdO
CdO

MoO
MoO,
ZrO
MoO
MoO,
ZrO
MoO
MoO
ZrO
MoO,
ZrO
ArO
CIO
SO

CaO
TiO, P02
TiO


Hydroxyl


NiOH
SbOH
SnOH
SnOH
SnOH
SnOH
InOH
CdOH

MoOH
MoOH
MoOH

MoOH


ZrOH
C10H
ArOH

C10H
CaOH
TiOH
TiOH


Nitrogen
AgN
AgN
NiN




SnN
SnN
SnN, CdN

MoN
MoN
MoN
MoN, ZrN


MoN, ZrN
MoN, ZrN

KN
ArN
ArN, CaN
ScN
TiN
VN


Chlorine

SrCl
ArCl

MoCl

MoCl
MoCl
MoCl
MoCl

SeCl
SeCl, AsCl
GeCl
GeCl, AsCl
SeCl, AsCl


GeCl

NCI, OC1


MgCl
SiCl, MgCl
SiCl


Sulfur

ZrS
CaS





MoS
MoS

SeS
SeS

SeS


GeS
SeS, GeS


SO

A1S
PS
S2, SO2H


Carbon
AgC
CdC
CuC


SnC

SnC
SnC
SnC


MoC

MoC


MoC, ZrC
yioC, ZrC
ArC
KC
ArC
CaC
TiC
VC
CrC


Other





















Mo++

Sn++
ArNa



                                   D-29/ICP-MS
ILM05.2

-------
Exhibit D  (ICP-MS) — Section 17
Tables/Diagrams/Flowcharts  (Con't)
Table 1.
Isobaric Molecular-Ion Interferences (Con't)
Analyte
207 Pb
204 Pb
55Mn
202Hg
200Hg
l"Hg
201Hg
198Hg
204Hg
196Hg
58Ni
60Ni
62Ni
61Ni
64Ni
80Se
78Se
82Se
76Se
"Se
74Se
107Ag
109Ag
205rŁ.^
203T1
51V
50V
64Zn
66Zn
68Zn
Oxygen


KO
WO
WO
WO

WO


CaO
CaO
TiO
ScO
TiO
ZnO
NiO
ZnO
NiO
NiO
NiO
ZrO



CIO
SO
TiO
TiO
CrO
Hydroxyl


ArOH

WOH
WOH
WOH
TaOH


KOH
CaOH
ScOH
CaOH
TiOH
CuOH
NiOH
CuOH
CoOH
NiOH
FeOH
ZrOH
MoOH

WOH
SOH

TiOH
TiOH
VOH
Nitrogen


KN

WN


WN

WN
CaN
TiN
TiN
TiN
TiN, CrN
ZnN
ZnN
ZnN
NiN
CuN
NiN

MoN


C1N
ArN
TiN, CrN
CrN
FeN
Chlorine










NaCl
MgCl, NaCl
A1C1, MgCl
MgCl
SiCl, A1C1
ScCl, CaCl
CaCl, KC1
TiCl, ScCl
KC1
CaCl, ArCl
C12, KC1
GeCl
GeCl


CIO, C1N

SiCl, A1C1
PCI, SiCl
PCI
Sulfur


NaS







MgS
SiS
SiS
SiS
S 2
ris
TiS
IiS, CrS
CaS
ScS
CaS
AsS
SeS


FS

S 2
S2
ArS
Carbon


CaC




we

we
TiC
TiC
TiC, CrC
TiC
CrC
ZnC
ZnC

ZnC
CuC
NiC
MoC
MoC


KC
ArC
CrC
FeC
FeC
Other


Cd++







Cd*+, Sn++
Sn t+
3n ++
Sn++












Mo++


Ba++
 ILM05.2
                       D-30/ICP-MS

-------
                                               Exhibit D  (ICP-MS) — Section 17
                                             Tables/Diagrams/Flowcharts  (Con't)
Table 1.
Isobaric Molecular-Ion Interferences (Con't)
Analyte
"Zn
70Zn
Oxygen
VO
FeO
Hydroxyl | Nitrogen | Chlorine
TiOH
CrOH
CrN
GeN
SCI
C12
Sulfur
CIS
ArS
Carbon
MnC
NiC
Other
Ba++

NOTE: The information provided in this table does not indicate that all of the
described interferences need to be tested.  However, this table can be
consulted if unusual samples are encountered.
                                  D-31/ICP-MS
                                                            ILM05.2

-------
Exhibit D (ICP-MS) - Section 17
Tables/Diagrams/Flowcharts  (Con't)

Table 2.    Mass Choices for Elements that Must Be Monitored
            During the Analytical Run
Mass
27.
121


15.
134,
135, 136, 137
1
111.
12,
114
53
59
63,
206,

21,
65
207. 208
25_, 26.
55
60.,
77,
107,

203,
61, 62
78, 80, 82.
109
205
51
66,
67, 68
Element of Interest
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Chromium
Cobalt
Copper
Lead
Magnesium
Manganese
Nickel
Selenium
Silver
Thallium
Vanadium
Zinc
 NOTE:  Underlined isotopes  are preferred  for measurements.  Where possible,
 alternative  isotopes  are indicated.   Those isotopes  not  listed shall  not be
 used as  a  primary isotope  for measurement, although  they may be monitored  for
 interference corrections if necessary.
 ILM05.2
D-32/ICP-MS

-------
                                               Exhibit D  (ICP-MS) — Section 17
                                             Tables/Diagrams/Flowcharts  (Con't)

Table 3.    Recommended Elemental Expressions for Isobaric Interferences
Element
Al
Sb
As
Ba
Be
Cd
Cr
Co
Cu
Pb
Mg
Mn
Ni
Se
Ag
Tl
V
Zn
Sc
Y
Rh
In
Tb
Ho
Bi
Isobaric
Correction
none
none
ArCl, Se
none
none
MoO, Pd
none
none
none
none
none
none
none
none
none
none
CIO, Cr
none
none
none
none
Sn
none
none
none
Expression Proportional to Elemental
Concentration
(1.0000) (27C)
(1.0000) (121C)
(1.0000) ( 75C) - (3.127) [("C) - (0.815) (82C)]
(1.0000) (137C)
(1.0000) (9C)
(1.000) ( 11LC) - (1.073) [ (108C) - (0.712) (106C) ]
(1.0000) (52C)
(1.0000) (59C)
(1.0000) (63C)
(1.0000) (206C) + (1.0000) (207C) + (1.0000) (208C)
(1.0000) (25C)
(1.0000) (55C)
(1.0000) (60C)
(1.0000) (78C)
(1.0000) (107C)
(1.0000) (205C)
(1.0000) (S1C) - (3.127) [ (53C) - (0.113) (" C) ]
(1.0000) (66C)
(1.0000) (45C)
(1.0000) (89C)
(1.0000) (103C)
(1.0000) (115C) - (0.0140) (118C)
(1.0000) (159C)
(1.0000) (165C)
(1.0000) (209C)
C - Calibration blank subtracted counts at specified mass

The coefficients in correction equations were calculated using natural
isotopic abundances, and assuming zero instrumental fractionation.  For each
particular instrument these coefficients must be determined experimentally.

The correction equations shall not be applied if appropriate interference
check sample measurement demonstrates absence of interference above the CRQL.
                                  D-33/ICP-MS
ILM05.2

-------
Exhibit D  (ICP-MS) - Section 17
Tables/Diagrams/Flowcharts  (Con't)
Table 4.
Internal Standards (must use at least three)
Internal Standard
Lithium
Scandium
Yttrium
Rhodium
Indium
Terbium
Holmium
Lutetium
Bismuth
Mass
6
45
89
103
115
159
165
175
209
CAS Number
7439-93-2
7440-20-2
7440-65-5
7440-16-6
7440-74-6
7440-27-9
7440-60-0
7439-94-3
7440-69-9
NOTE: Use  of  Li6 requires enriched standard.
Table  5.
Spiking Levels for Spike Sampl"e Analysis
Analyte
Al
Sb
As
Ba
Be
Cd
Cr
Co
Cu
Pb
Mn
Ni
Se
Ag
Tl
V
Zn
Spike (ug/L)
2000
100
40
2000
50
50
200
500
250
20
500
500
10
50
50
500
500
 ILM05.2
                       D-34/ICP-MS

-------
     EXHIBIT  D  -  PART  C

     ANALYTICAL METHODS
            FOR
COLD VAPOR MERCURY ANALYSIS
        D-1/Mercury                           ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                           D-2/Mercury

-------
        Exhibit D - Analytical Methods for Cold Vapor Mercury Analysis

                               Table of Contents

Section                                                                     Page

1.0   SCOPE AND APPLICATION 	   5

2.0   SUMMARY OF METHOD	5
      2.1   Water by Automated and Manual Techniques   	  5
      2.2   Soil/Sediment by Manual Technique  	   5

3.0   DEFINITIONS 	   6

4.0   INTERFERENCES 	   6
      4.1   Water	6
      4.2   Soil/Sediment 	   6

5.0   SAFETY   	7

6.0   EQUIPMENT AND SUPPLIES   	  7
      6.1   General Information for Water and  Soils  (Automated and
            Manual Techniques)   	  7
      6.2   Water by Automated Technique   	  7
      6.3   Water and Soil/Sediment by Manual  Techniqu e  	  7

7.0   REAGENTS AND STANDARDS   	  8
      7.1   Reagents   	8
      7.2   Standards	9

8.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE    	  10
      8.1   Sample Collection and Preservation  	  10
      8.2   Procedure for Sample Storage   	  10
      8.3   Contract Required Holding Time   	  10

9.0   CALIBRATION AND STANDARDIZATION	11
      9.1   Cold Vapor Atomic Absorption  (AA)  Instrument
            Calibration Procedure 	 11
      9.2   Initial Calibration Verification  (ICV)    	  11
      9.3   Continuing Calibration Verification  (CCV  ) 	  11
      9.4   Initial and Continuing Calibration Blank  (ICB/CCB)    	  12

10.0  PROCEDURE	13
      10.1  Sample Preparation   	 13
      10.2  Sample Analysis	15

11.0  DATA ANALYSIS AND CALCULATIONS   	17
      11.1  Water/Aqueous by Automated Technique   	  17
      11.2  Water/Aqueous by Manual Technique  	 17
      11.3  Soil by Manual Technique   	17
      11.4  Adjusted Method Detection Limit  (MDL)/Adjusted Contract
            Required Quantitation Limit  (CRQL) Calculation    	  17

12.0  QUALITY CONTROL	19
      12.1  Initial Calibration Verification  (ICV)    	  19
      12.2  Continuing Calibration Verification  (CCV  ) 	  19
      12.3  Contract Required Quantitation Limit  (CRQL) Check
            Standard  (CRI)   	19
      12.4  Blank Analyses   	19
      12.5  Spike Sample Analysi s	20
      12.6  Duplicate Sample Analysi s 	 21
      12.7  Laboratory Control Sample (LCS) Analysis   	  22
      12.8  Method Detection Limit  (MDL) Determination    	  23

13.0  METHOD PERFORMANCE   	 24


                                  D-3/Mercury                           ILM05.2

-------
        Exhibit D - Analytical Methods for Cold Vapor Mercury Analysis

                           Table of Contents (Con't)
Section                                                                     Page

14.0  POLLUTION PREVENTION   	 24

15.0  WASTE MANAGEMENT   	 24

16.0  REFERENCES   	24

17.0  TABLES/DIAGRAMS/FLOWCHARTS   	 24
 ILM05.2                           D-4/Mercury

-------
                                          Exhibit D (Mercury)  — Sections 1 & 2
                                                          Scope and Application
1.0
SCOPE AND APPLICATION
      The analytical method that follows is designed to analyze water,
      sediment, sludge, and soil samples taken from hazardous waste sites
      using a cold vapor technique with Atomic Absorption  (AA) for total
      mercury.

      In addition to inorganic forms of mercury, organic mercurials may also
      be present.  These organo-mercury compounds will not respond to the cold
      vapor AA technique unless they are first broken down and converted to
      mercuric ions.  Potassium permanganate oxidizes many of these compounds,
      but studies have shown that a number of organic mercurials, including
      phenyl mercuric acetate and methyl mercuric chloride, are only partially
      oxidized by this reagent.  Potassium persulfate has been found to give
      approximately 100% recovery when used as the oxidant with these
      compounds.  Therefore, a persulfate oxidation step following the
      addition of the permanganate has been included to ensure that organo-
      mercury compounds, if present, will be oxidized to the mercuric ion
      before measurement.  A'heat step is required for methyl mercuric
      chloride when present in, or spiked to, a natural system.

      The range of the method may be varied through instrument and/or recorder
      expansion.  Using a 100 milliliters  (mL) sample, a detection limit of
      less than 0.1 micrograms per Liter  (pg/L) can be achieved.

      The range of the method for soil/sediments is 0.05 milligrams per
      kilogram  (mg/kg) to 5 mg/kg.  The range may be extended above or below
      the normal range by increasing or decreasing sample size or through
      instrument and recorder control.

2.0   SUMMARY OF METHOD

2.1   Water by Automated and Manual Techniques

      This is a physical method based on the absorption of radiation at 253.7
      nanometers  (nm) by mercury vapor.  Free mercury atoms can exist at room
      temperature; therefore, mercury can be measured by Atomic Absorption
       (AA) without a heated sample cell.  Organic compounds are oxidized, and
      in the cold vapor mercury technique, mercury is chemically reduced to
      the free atomic state by reacting the sample with a strong reducing
      agent like stannous chloride or sodium borohydride in a closed reaction
      vessel.  The volatile free mercury is then driven from the reaction
      flask by bubbling air through the solution.  Mercury atoms are carried
      in the air stream through tubing connected to an absorption cell, which
      is placed in the light path of the AA spectrophotometer.  Sometimes the
      cell is heated slightly to avoid water condensation; otherwise the cell
      is completely unheated.  As the mercury atoms pass into the sampling
      cell, measured absorbance rises indicating the increasing concentration
      of mercury atoms in the light path.  Some systems allow the mercury
      vapor to pass from the absorption tube to waste, in which case the
      absorption peaks and then falls as the mercury is depleted.  The highest
      absorbance observed during the measurement will be taken as the
      analytical signal.

2.2   Soil/Sediment by Manual Technique

2.2.1    A weighed portion of the sample is acid digested for 2 minutes at
         95°C,  followed by oxidation with potassium permanganate and potassium
         persulfate.  Mercury in the digested sample is then measured by the
         conventional cold vapor technique.

2.2.2    An alternate digestion involving the use of an autoclave is described
         in Section 10.1.4.2.1.2.
                                   D-5/Mercury
                                                                   ILM05.2

-------
Exhibit D (Mercury) — Sections 3 & 4
Definitions

3.0   DEFINITIONS

      See Exhibit  G for a complete list of definitions.

4.0   INTERFERENCES

4.1   Water

4.1.1    Some sea  waters and wastewaters high in chlorides have shown a
         positive  interference, and require additional permanganate  [as much
         as 25 milliliters  (mL)].  During the oxidation step, chlorides are
         converted to  free  chlorine which will also absorb radiation at 253
         nanometers  (nm).   Care must be taken to assure that free chlorine is
         absent before the  mercury is reduced and  swept into the cell.  This
         may be accomplished by using an excess of hydroxylamine sulfate
         reagent  (25 mL).   Both inorganic and organic mercury spikes have been
         quantitatively recovered from the sea water using this technique.

4.1.2    Formation of  a heavy precipitate, in some wastewaters and effluents,
         has been  reported  upon addition of concentrated  sulfuric acid.  If
         this is encountered, the problem sample cannot be analyzed by this
         method.

4.1.3    Possible  interference  from sulfide is eliminated by the addition of
         potassium permanganate.  Concentrations as high  as 20 milligram per
         Liter  (mg/L)  of sulfide as sodium sulfide do not interfere with the
         recovery  of added  inorganic mercury from  reagent water.

4.1.4    Copper has also been reported to interfere; however, copper
         concentrations as  high as 10 mg/L have no effect on recovery of
         mercury  from  spiked  samples.

4.1.5    Samples containing solids must be blended and then mixed while being
         sampled  if  total mercury values are to be reported.

4.2    Soil/Sediment

4.2.1    The same  types of  interferences that may  occur in water samples are
         also possible with sediments  (i.e., sulfides, high copper,  high
         chlorides,  etc.).

4.2.2    Samples containing high concentrations of oxidizable organic
         materials,  as evidenced by high chemical  oxygen  demand values, may
         not be  completely  oxidized by this procedure.  When this occurs, the
         recovery  of  organic  mercury will be low.  The problem can be
         eliminated  by reducing the weight of  the  original sample or by
         increasing  the amount  of potassium persulfate  (and consequently
         stannous  chloride) used in  the digestion.
 ILM05.2                           D-6/Mercury

-------
                                          Exhibit D (Mercury) — Sections 5 & 6
                                                                          Safety

5.0   SAFETY

      See Section 1.14 in Exhibit D - Introduction to Analytical Methods.

6.0   EQUIPMENT AND SUPPLIES

      Brand names, suppliers, and part numbers are for illustrative purposes
      only.  No endorsement is implied.  Equivalent performance may be
      achieved using equipment and supplies other than those specified here,
      however, a demonstration of equivalent performance meeting the
      requirements of this Statement of Work  (SOW) is the responsibility of
      the Contractor.  The Contractor shall document any use of alternate
      equipment or supplies in the Sample Delivery Group  (SDG) Narrative.

6.1   General Information for Water and Soils  (Automated and Manual
      Techniques)

6.1.1    Atomic Absorption  (AA) Spectrophotometer - Any AA unit having an open
         sample presentation area in which to mount the absorption cell is
         suitable.  Instrument settings recommended by the particular
         manufacturer should be followed.

         NOTE: Instruments designed specifically for the measurement of
         mercury using the cold vapor technique are commercially available and
         may be substituted for the AA spectrophotometer.

         NOTE: All cold vapor mercury analyzers shall be equipped with all
         manufactured required equipment  (i.e., dryers) to ensure that the
         specified CRQLs are met.

6.1.2    Mercury Hollow Cathode Lamp

6.1.3<'-*  Recorder - Any multi-range variable speed recorder  that is compatible
         with the UV detection system is suitable.

6.2   Water by Automated Technique

6.2.1    Automated Analyzer instrumentation consisting of:

6.2.1.1     Sampler with provision for sample mixing

6.2.1.2     Manifold

6.2.1.3     Proportioning Pump(s)

6.2.1.4     High temperature heating bath with distillation  coil(s)

6.2.1.5     Vapor-liquid separator

6.2.1.6     Absorption cell with quartz windows

6.3   Water and Soil/Sediment by Manual Technique

6.3.1    Absorption Cell - Standard spectrophotometer cells

6.3.2    Air Pump - Any device capable of delivering 1 Liter' (L) of air per
         minute may be used.

6.3.3    Flowmeter - Capable of measuring an air flow of 1 L per minute.

6.3.4    Aeration Tubing - Tygon tubing is used for transporting the mercury
         vapor from the sample bottle to the absorption cell and for its
         return.

                                  D-7/Mercury       -,                    ILM05.2

-------
Exhibit D (Mercury) — Sections 6 & 1
Reagents and Standards

6.3.4.1     Straight glass tubing terminating in a coarse porous frit is used
            for sparging air  into the sample.

6.3.5    Drying Tube - 6" X 3/4" diameter tube containing 20 grams  (g) of
         magnesium perchlorate.

         NOTE: In  place of the magnesium perchlorate drying tube, a small
         reading lamp with a  60-watt bulb may be used to prevent condensation
         of moisture inside the cell.  The lamp is positioned  to shine on the
         absorption cell maintaining the air temperature in the cell  about
         10°C above ambient temperature.

7.0   REAGENTS AND STANDARDS

7.1   Reagents

7.1.1    Water by  Automated Technique

7.1.1.1     Reagent Water - The purity of this water must be equivalent to
            ASTM Type  II water  (ASTM D1193-77).  Use this preparation for all
            reagents,  standards, and dilutions of  solutions.

7.1.1.2     Sulfuric acid, concentrated - Reagent  grade.

7.1.1.2.1      Sulfuric acid, 2N -  Dilute 56 milliliters  (mL)  of concentrated
                sulfuric  acid to  1  Liter  (L)  with reagent  water.

7.1.1.2.2      Sulfuric acid, 10%  - Dilute 100 mL  concentrated sulfuric acid
                to  1  L  with  reagent water.

7.1.1.3     Nitric acid, concentrated - Reagent grade of  low mercury  content.

            Nitric acid,  0.5% wash solution  -  Dilute  5  mL of  concentrated
            nitric acid  to  1  L with reagent  water.

7.1.1.4     Stannous sulfate  - Add 50 grams  (g) stannous  sulfate to 500 mL  of
            2N  sulfuric  acid (see  Section  7.1.1.2.1).   This mixture is a
            suspension and  should  be stirred continuously during use.

            NOTE:  Stannous  chloride may be used  in place  of stannous  sulfate.

7.1.1.5     Sodium chloride-hydroxylamine sulfate  solution  -  Dissolve 30 g  of
            sodium chloride  and  30 g of hydroxylamine  sulfate in reagent water
            to  1  L.

            NOTE:  Hydroxylamine  hydrochloride may  be  used in  place of
            hydroxylamine sulfate.

7.1.1.6     Potassium  permanganate (KMnO4) - 0.5% solution, w/v.  Dissolve  5 g
            of  potassium permanganate  in  1  L of  reagent water.

7.1.1.7     Potassium  permanganate, 0.1N  - Dissolve 3.16  g of potassium
            permanganate in reagent water and dilute  to 1 L.

7.1.1.8     Potassium  persulfate - 0.5% solution,  w/v.   Dissolve 5 g  of
             potassium persulfate in 1  L of  reagent water.

7.1.1.9     Air  scrubber solution  - Mix equal  volumes  of  0.1N potassium
             permanganate (see Section 7.1.1.6)  and 10% sulfuric acid  (see
             Section 7.1.1.2.2).
 ILM05.2                           D-8/Mercury

-------
                                               Exhibit D (Mercury) — Section 7
                                                 Reagents and Standards  (Con't)

7.1.2    Water and Soil/Sediment by Manual Technique

7.1.2.1     Reagent water - The purity of this water must be equivalent to
            ASTM Type II water  (ASTM Dl193-77).  Use this preparation for all
            reagents, standards, and dilutions of  solutions.

7.1.2.2     Sulfuric acid, concentrated - Reagent  grade.

7.1.2.2.1      Sulfuric acid, 0.5N - Dilute 14.0 mL of concentrated sulfuric
               acid to  1 L.  (Water technique  only.)

7.1.2.3     Nitric acid, concentrated - Reagent grade of low mercury content.
            If a high Preparation Blank  (PB)  is obtained, it may  be necessary
            to distill  the nitric acid. .

7.1.2.4     Stannous sulfate - Add 25 g stannous sulfate to 250 mL of 0.5N
            sulfuric acid.  This mixture is a suspension and should be  stirred
            continuously during use.

            NOTE: Stannous chloride may be used in place of stannous sulfate.

7.1.2.5     Sodium chloride-hydroxylamine sulfate  solution - Dissolve 12 g of
            sodium chloride and 12 g of hydroxylamine sulfate in  reagent water
            and dilute  to 100 mL.

            NOTE: Hydroxylamine hydrochloride may  be used in place of
            hydroxylamine sulfate.

7.1.2.6     Potassium permanganate  (KMn04)  - 5% solution,  w/v.   Dissolve 5 g
            of potassium permanganate in 100  mL of reagent water.

7.1.2.7,    Potassium persulfate - 5% solution, w/v.  Dissolve 5  g of
      '-'-     potassium persulfate in 100 mL of reagent water.

7.2   Standards

7.2.1    Introduction

         The Contractor must provide all standards to be used with this
         contract.  These standards may be used only after they have been
         certified according to the procedure in Exhibit E, Section 8.0.  The
         Contractor must be able to verify that the standards are certified.
         Manufacturer's certificates of analysis must be retained by the
         Contractor and presented upon request.

7.2.1.1     Stock standard solutions may be purchased or prepared from ultra
            high purity grade chemicals or metals.

7.2.1.2     Stock mercury solution - Dissolve 0.1354 g of mercuric chloride in
            75 mL of reagent water.  Add 10 mL of  concentrated nitric acid and
            adjust the  volume to 100.0 mL  [1.0 mL  = 1.0 milligram (ing) Hg] .

7.2.1.3     Working mercury solution - Make successive dilutions  of the stock
            mercury solution  (see Section 7.2.1.2) to obtain a working
            standard containing 0.1 micrograms  (ug) per mL.  This working
            standard and the dilutions of the stock mercury solution should be
            prepared fresh daily.  Acidity of the  working standard should be
            maintained  at 0.15% nitric acid.   This acid should be added to the
            flask as needed before the addition of the aliquot.   From this
            solution, prepare standards.
                                   D-9/Mercury                            ILM05.2

-------
Exhibit D  (Mercury) — Sections 7 & 8
Sample Collection, Preservation, and Storage

7.2.2    Working  Standards

7.2.2.1     Contract Required Quantitation Limit  (CRQL) Check Standard  (CRI)

            The concentration of the CRI  for  mercury  shall be at  the  CRQL.
            Information  regarding the  CRI shall be  reported on  Form IIB-IN.

7.2.2.2     Method Detection Limit  (MDL)  Solution

            The MDL  solution shall  be  at  a concentration  of 3 to  5  times  the
            expected MDL.

8.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE

8 .1   Sample Collection  and Preservation

      All  samples must be  collected in glass  or polyethylene containers.
      Water/aqueous  samples must be preserved with  nitric acid  to pH  less than
      2  immediately  after  collection.  All samples  must be iced or
      refrigerated at 4°C  (ą2°C) from  the time of collection until  digestion.

8.1.1    Dissolved Metals

         For  the  determination  of dissolved metals,  the sample  must be
         filtered through  a  0.45 micrometer  (urn)  pore diameter  membrane filter
         at the  time of  collection  or  as  soon as  possible.  Use a portion of
         the  sample  to  rinse the filter  flask, discard this portion,  and
         collect  the required volume of  filtrate.   Preserve the filtrate with
         nitric  acid to  pH less than 2 immediately  after  filtration.

8.2   Procedure  for  Sample Storage

      The  samples must be  protected from light and  refrigerated at  4°C  (ą2°C)
      from the time  of  receipt  until 60  days  after  delivery of  a  complete,
      reconciled data package to USEPA.   After 60 days the samples  may  be
      disposed of in a manner that  complies with  all  applicable regulations.

8.3   Contract Required  Holding Time

      The  maximum holding  time  for  mercury is 26  days from Validated  Time of
      Sample  Receipt (VTSR).
 ILM05.2                          D-10/Mercury

-------
                                               Exhibit D (Mercury) — Section 9
                                                Calibration and Standardization
9.0   CALIBRATION AND STANDARDIZATION
9.1   Cold Vapor Atomic Absorption  (AA) Instrument Calibration Procedure

9.1.1    Instruments shall be calibrated daily or once every 24 hours and each
         time the instrument is set up.  The instrument standardization date
         and time shall be included in the raw data.

9.1.2    The date and time of preparation and analysis shall be given in the
         raw data.

9.1.3    Calibration standards shall be prepared fresh with each preparation
         batch.  Prepare a minimum of five calibration standards  (which
         includes a blank) in graduated amounts in the appropriate range.  One
         of the standards must be at the Contract Required Quantitation Limit
         (CRQL).

9.1.4    Aspirate the standards and record the readings.  Results for these
         standards shall be within 5% of the true value.  Each standard
         concentration and the calculations to show that the 5% criteria has
         been met shall be given in the raw data.  If the values do not fall
         within this range, recalibration is necessary.  The 5% criteria does
         not apply to the calibration standard at the CRQL.  The acceptance
         criteria for the initial calibration curve is a correlation
         coefficient more than or equal to 0.995.

9.1.5    Baseline correction is acceptable as long as it is performed after
         every sample or after the Continuing Calibration Verification  (CCV)
         and Blank  (CCB) check.  Resloping is acceptable as long as it is
         immediately preceded and immediately followed by a compliant CCV and
         CCB.

9.2   "^Initial Calibration Verification (ICV)

9.2.1 ••   Immediately after the AA system has been calibrated, the accuracy of
         the initial calibration shall be verified and documented for mercury
         by the analysis of the ICV solution at the wavelength used for
         analysis.

9.2.2    Only if the ICV solution is not available from USEPA, or where a
         certified solution of the analyte is not available from any source,
         analyses shall be conducted on an independent standard at a
         concentration other than that used for instrument calibration, but
         within the calibration range.  An independent standard is defined as
         a standard composed of the analyte from a different source than that
         used in the standards for the instrument calibration.  The value for
         the ICV shall be reported on Form IIA-IN.

9.3   Continuing Calibration Verification (CCV)

9.3.1    To ensure calibration accuracy during each analysis run, one of the
         following standards is to be used for the CCV and shall be analyzed
         and reported at a frequency of 10% or every 2 hours during an
         analysis run, whichever is more frequent.  The standard shall also be
         analyzed and reported at the beginning of the run and after the last
         analytical sample.  The analyte concentration in the CCV standard
         shall be different than the concentration used for the ICV and shall
         be one of the following solutions at or near the mid-range level of
         the calibration curve:
                                  D-ll/Mercury
ILM05.2

-------
Exhibit D (Mercury) — Section 9
Calibration and Standardization  (Con't)

         ^     USEPA  Solutions

         ••ť     NIST Standards

         '*:     A  Contractor-prepared  standard  solution

         The  same CCV standard shall  be used throughout the analysis runs  for
         a  Sample Delivery  Group (SDG) of samples  received.

9.3.2    Each CCV analyzed  shall reflect the conditions of analysis of all
         associated analytical samples  (the preceding 10 analytical samples or
         the  preceding analytical samples up to  the previous CCV).  The
         duration of  analysis, rinses, and other related operations that may
         affect the CCV measured result may not  be applied to  the CCV to a
         greater  extent than the extent applied  to the associated analytical
         samples. For instance,  the  difference  in time between  a CCV analysis
         and  the  blank immediately following it, as well as the  difference in
         time between the CCV and the analytical sample immediately preceding
         it,  may  not  exceed the  lowest difference  in time between any two
         consecutive  analytical  samples associated with the CCV.

9.3.3    Information  regarding the CCV shall be  reported on Form IIA-IN.

9.4    Initial and Continuing Calibration Blank (ICB/CCB)

       A calibration blank shall  be analyzed at each wavelength used for
       analysis immediately  after every ICV and CCV, at a frequency of 10%  or
       every 2 hours during  the run, whichever  is more frequent.  The blank
       shall be analyzed at  the beginning of the  run and after  the last
       analytical  sample.

       NOTE: A CCB shall be  analyzed immediately  after the last CCV, and the
       last  CCV shall  be analyzed immediately after the last analytical sample
       of the  run. The results for the calibration blanks shall  be reported on
       Form  III-IN.
 ILM05.2                          D-12/Mercury

-------
                                              Exhibit D  (Mercury) — Section 10
                                                                       Procedure

10.0  PROCEDURE

10.1  Sample Preparation

10.1.1   If insufficient sample amount  (less than 90% of the required amount)
         is received to perform the analyses, the Contractor shall  contact the
         Sample Management Office  (SMO) to inform them of  the problem.  SMO
         will contact the Region for instructions.  The  Region will either
         require that no sample analyses be performed or will require that a
         reduced volume be used for the sample analysis.   No other  changes in
         the analyses will be permitted.  The Contractor shall document the
         Region's decision in the  Sample Delivery Group  (SDG) Narrative.

10.1.2   If multiphase samples  (e.g., two-phase liquid sample, oily
         sludge/sandy soil sample) are received by the Contractor,  the
         Contractor shall contact  SMO to apprise them of the type of sample
         received.  SMO will contact the Region.  If all phases  of  the sample
         are amenable to analysis, the Region may require  the Contractor  to do
         any of the following:

         :*:    Mix the sample and  analyze an aliquot from  the homogenized
               sample.

         ==    Separate the phases of the sample, and analyze one or more of
               the phases separately.   SMO will provide  EPA sample  numbers for
               the additional phases, if required.

         :i:    Do not analyze the  sample.

10.1.2.1    If all of the phases are not amenable to analysis  (i.e., outside
            the scope), the Region may require the Contractor to do any of the
            following:

            :x'    Separate  the  phases and analyze the phase (s)  that is  (are)
                  amenable  to analysis.  SMO will provide  EPA  sample numbers
                  for the additional phases, if required.

            '•*'•    Do not analyze  the sample.

10.1.2.2    No other changes in the analyses will be permitted.  The
            Contractor shall document the Region's decision in the  SDG
            Narrative.

10.1.3   Water Preparation of Standards and Samples  (Manual Technique)

10.1.3.1    Standards Preparation

10.1.3.1.1     Transfer aliquots of the working mercury  solution to a series
               of 300 milliliters  (mL)  BOD bottles.  Add enough  reagent water
               to each bottle to make a total volume of  100 mL.

10.1.3.1.2     Mix thoroughly and  add 5 mL of concentrated sulfuric acid  (see
               Section 7.1.2.2) and 2.5 mL of concentrated nitric acid  (see
               Section 7.1.2.3) to each bottle.  Add 15  mL of KMnO  4 (see
               Section 7.1.2.6) solution to each bottle  and allow to stand at
               least 15 minutes.   Add 8 mL of potassium  persulfate  (see
               Section 7.1.2.7) to each bottle and heat  for 2 hours in a  water
               bath maintained  at  95°C.   (If an autoclave  is employed, cover
               the BOD bottles  with foil and heat in the autoclave  for 15
               minutes at 120°C and 15  PSI instead of heating for 2 hours in a
               waterbath at 95°C).  Cool and add 6 mL of sodium  chloride-
               hydroxylamine sulfate solution  (see Section 7.1.2.5) to reduce
               the excess permanganate.  When the solution has been

                                  D-13/Mercury                          ILM05.2

-------
Exhibit D  (Mercury) — Section 10
Procedure  (Con't)

                decolorized,  wait  30  seconds,  add 5  mL  of  the  stannous sulfate
                solution  (see Section 7.1.2.4)  and immediately attach the
                bottle  to the aeration apparatus to  form a closed system.  At
                this point the sample is  allowed to  stand  quietly without
                manual  agitation.

10.1.3.1.3      The circulating pump,  which  has  previously been adjusted to a
                rate of 1 Liter (L) per minute,  is allowed to  run continuously
                (see Note 1).  The absorbance will increase and reach maximum
                within  30 seconds. As soon  as the response levels off,  open
                the bypass valve  and  continue the aeration until the absorbance
                returns to its minimum value (see Note  2) .  Close the bypass
                valve,  remove the stopper and frit from the BOD bottle and
                continue  the aeration.  Proceed with the standards and
                construct a standard  curve by plotting  instrument response at
                253 nanometers (nm) versus micrograms (ug) of  mercury.

                NOTE  1: An open system where the mercury vapor is passed
                through the absorption cell only once may  be used instead of
                the closed system.

                NOTE  2: Because of the toxic nature  of  mercury vapor,
                precaution must be taken to avoid its inhalation.  Therefore, a
                bypass  has been included in the system  to  either vent the
                mercury vapor into an exhaust hood or pass the vapor through
                some  absorbing media, such as equal  volumes of 0.1 M KMnO4,  and
                10% H2S04 or  0.25% iodine in a 3% KI solution.   A  specially
                treated charcoal that will adsorb mercury  vapor is commercially
                available.

 10.1.3.2     Sample Preparation

 10.1.3.2.1     Preparation Method/Code (CW1)

 10.1.3.2.1.1      Transfer 100 mL, or an aliquot diluted  to 100 mL, containing
                   not more  than  1.0  ug of mercury,  to a  300 mL BOD bottle and
                   continue  as described in Section 10.1.3.1.2.

                   NOTE: The same amount of KMn04 added to the samples  should
                   be present in  standards and blanks.

 10.1.3.2.1.2      Cool and add 6 mL  of sodium chloride-hydroxylamine sulfate
                   (see Section 7.1.2.5)  to reduce the excess permanganate.
                   Purge the headspace in the BOD bottle  for at least 1  minute
                   and add 5 mL of stannous sulfate  (see  Section  7.1.2.4) and
                   immediately attach the bottle to the aeration  apparatus.

                   NOTE: Add reductant in 6 mL  increments  until KMnO  4 is
                   completely reduced  (until the color is  no longer purple).

 10.1.4   Soil/Sediment Preparation of Standards and Samples (Manual)

 10.1.4.1     Standards  Preparation

 10.1.4.1.1     Transfer aliquots of  the working mercury solutions  (see  Section
                7.2.1.3) to  a series of 300 mL  BOD bottles.  Add  enough  reagent
                water  to each bottle to make a  total volume of  10 mL.

 10.1.4.1.2     Add 5 mL of  concentrated H2SO4   (see Section 7.1.2.2) and 2.5 mL
                of concentrated HN03  (see Section 7.1.2.3) and heat 2 minutes
                in a water bath at 95°C.  Allow  the sample to  cool  and  add  50
                mL reagent water, 15 mL of KMn04  solution (see Section  7.1.2.6)
                and 8  mL of  potassium persulfate solution (see  Section  7.1.2.7)

 ILM05.2   )                        D-14/Mercury

-------
                                              Exhibit D (Mercury)  — Section 10
                                                              Procedure  (Con't)

               to each bottle and return to the water bath for 30 minutes.
               Cool and add 6 mL of sodium chloride-hydroxylamine sulfate
               solution  (see Section 7.1.2.5) to reduce the excess
               permanganate.  Add 50 mL of reagent water  (final volume of
               reagent water = 100 mL).  Treating each bottle individually,
               add 5 mL of stannous sulfate solution  (see Section 7.1.2.4) and
               immediately attach the bottle to the aeration apparatus.  At
               this point the sample is allowed to stand quietly without
               manual agitation.  If an autoclave is used, the standards shall
               be prepared in the same way as the samples  (see Section
               10.1.4.2.1.2).

10.1.4.1.3     The circulating pump,  which has previously been adjusted to a
               rate of 1 L per minute, is allowed to run continuously.  The
               absorbance, as exhibited either on the spectrophotometer or the
               recorder, will increase and reach maximum within 30  seconds.
               As soon as the response levels off, open the bypass  valve and
               continue the aeration until the absorbance returns to its
               minimum value.  Close the bypass valve, remove the fritted
               tubing from the BOD bottle and continue the aeration.  Proceed
               with the standards and construct a standard curve by plotting
               peak height versus ug of mercury.

10.1.4.2    Sample Preparation

10.1.4.2.1     Preparation Method/Code (CS1)

10.1.4.2.1.1      Weigh a representative 0.20 g (ą0.01 g)  portion of wet
                  sample and place in  the bottom of a BOD bottle.   Add enough
                  reagent water to each sample to make a total volume of 10
                  mL.  Continue as described in Section 10.1.4.1.2.

10.1.4.2.1.2      If an autoclave is used, add 5 mL of concentrated H 2S04  and
                  2 mL of concentrated HN03 to the 0.20 g (ą0.01 g)   of
                  sample.  Add 5 mL of saturated KMn04 solution and 8 mL of
                  potassium persulfate solution and cover with a piece of
                  aluminum foil.  The  sample is autoclaved at 120°C and  15 PSI
                  for 15 minutes.  Cool, make up to a volume of 100 mL with
                  reagent water, and add  6 mL of sodium chloride-hydroxylamine
                  sulfate solution  (see Section 7.1.2.5) to reduce  the excess
                  permanganate.  Purge the headspace of the sample  bottle for
                  at least one minute  and continue as described under Section
                  10.1.4.1.2.

10.1.5   Preparation of Standards for Automated Cold Vapor Analysis Technique
          (Analysis Method - AV)

10.1.5.1    Standards Preparation

            Make successive dilutions of the stock mercury solution to obtain
            a working standard containing 0.1 ug per mL.  This working
            standard and the dilutions of the stock mercury solution should be
            prepared fresh daily.  Acidity of the working standard  should be
            maintained at 0.15% nitric acid.  This acid should be added to the
            flask as needed before the addition of the aliquot.  From this
            solution, prepare standards.

10.2  Sample Analysis

10.2.1   Set up instrument with proper operating parameters.

10.2.2   Profile and calibrate instrument according to instrument
         manufacturer's recommended procedures, using calibration standard

                                  D-15/Mercury                          ILM05.2

-------
Exhibit D (Mercury)  — Section 10
Procedure (Con't)

         solutions mentioned in Section 9.1.  Samples prepared by a certain
         method must be analyzed with calibration and QC standards prepared by
         the same method.  Therefore, only one Preparation Method/Code can be
         associated with each run.

10.2.3   Analyze the Continuing Calibration Verification (CCV) instrument
         check standard and the Continuing Calibration Blank  (CCB) after every
         10 analytical samples.

10.2.4   Analysis of Water/Aqueous Samples by the Automated Cold Vapor
         Technique  (AV) Preparation Method/Code  (CW2)

10.2.4.1    Set up manifold.

10.2.4.2    Feed all the reagents through the system with acid wash solution
             (see Section 7.1.1.3) through the sample line, adjusting  the
            heating bath to 105°C.

10.2.4.3    Turn on the Atomic Absorption (AA) Spectrophotometer, adjust
            instrument settings as recommended by the manufacturer, align
            absorption cell in light path for maximum transmittance and place
            heat lamp directly over absorption cell.

10.2.4.4    Arrange working mercury standards in sampler and  start sampling.
            Complete  loading  of sample  tray  with unknown samples.

10.2.4.5    After the analysis is complete,  put all lines except the  H  2SOt
            line in reagent water to wash out system.  After  flushing, wash
            out  the H2S04 line.  Also flush the coils in the high temperature
            heating bath by pumping  stannous sulfate  (see Section  7.1.1.4)
            through the  sample lines followed by reagent water.  This will
            prevent build-up  of oxides  of manganese.
 ILM05.2                          D-16/Mercury

-------
                                              Exhibit D (Mercury) — Section 11
                                                 Data Analysis and Calculations

11.0  DATA ANALYSIS AND CALCULATIONS

11.1  Water/Aqueous by Automated Technique

11.1.1   Prepare a standard curve by plotting the instrumental response of
         processed standards against true concentration values.  Use a linear
         regression equation to determine the concentration of field and
         Quality Control  (QC) samples.

11.1.2   If samples were diluted for analysis, multiply the results from the
         linear regression by the dilution  factor.

11.2  Water/Aqueous by Manual Technique

11.2.1   Determine the instrumental response of the unknown and determine the
         mercury value from the standard curve.

11.2.2   Calculate the mercury concentration in the sample by the formula:

         EQ. 1  Aqueous Sample Concentration  (Manual)

             Hg Concentration (pg/L)  =  	ug Hg, curve	 x 1000 mL
                                       aliquot volume, mL     1 L


11.3  Soil by Manual Technique

11.3.1   Measure the instrumental response  of the unknown and determine the
         mercury value from the standard curve.

11.3.2   Calculate the mercury concentration in the sample by the formula:

         EQ. 2  Soil Sample Concentration  (Manual)


               Hg Concentration (mg/kg) = Hg pg/g  = 	 x (0.1L)
                                                    W X S
          WHERE,   C    =   Concentration from curve (ug/L)

                   W    =   Wet sample weight (g)

                   S    =   % Solids/100 (see Exhibit D - Introduction to
                            Analytical Methods,  Section 1.6).


11.4  Adjusted Method Detection Limit  (MDL)/Adjusted Contract Required
      Quantitation Limit  (CRQL) Calculation

      To calculate the adjusted MDL or adjusted CRQL for water/aqueous
      samples, multiply the value of the MDL  (pg/L) or CRQL  (ug/L) by the
      Dilution Factor.  Calculate the adjusted MDL or adjusted CRQL for  soil
      samples as follows:
                                  D-17/Mercury                          ILM05.2

-------
Exhibit D  (Mercury) — Section 11
Data Analysis and Calculations (Con't)
      EQ. 3  Adjusted Soil MDL/Adjusted Soil CRQL Concentration

                                                          W    i
            Adjusted Concentration (dry wt.) (mg/kg)  = C x  — x — x DF
                                                          WR   S
       WHERE,       C   =   MDL or CRQL concentration (mg/kg)

                    WM   =   Method required wet sample weight  (g)

                    WR   =   Reported wet sample weight  (g)

                    S   =   % Solids/100 (see Exhibit D - Introduction to
                            Analytical Methods, Section 1.6).

                    DF  =   Dilution Factor
 ILM05.2
D-18/Mercury

-------
                                              Exhibit D (Mercury) — Section 12
                                                                 Quality Control

12.0  QUALITY CONTROL

12.1  Initial Calibration Verification (ICV)

      The ICV Standard shall be prepared in the same acid matrix as the
      samples and carried through the entire preparation and analysis
      procedure.  If measurements exceed the control limits of 80%  (low)  and
      120% (high), the analysis shall be terminated, the problem corrected,
      the instrument recalibrated, and the calibration reverified.
      Information regarding the ICV shall be reported on Form IIA-IN.

12.2  Continuing Calibration Verification (CCV)

      The CCV Standard shall be prepared by the analyst at a concentration
      equivalent to the mid-point of the calibration curve and carried through
      the entire preparation and analysis procedure.  If the deviation of the
      CCV is greater than the control limits of 80%  (low) and 120%  (high), the
      analysis shall be stopped, the problem corrected, the instrument
      recalibrated, the calibration verified, and re-analysis of the preceding
      10 analytical samples or all analytical samples analyzed since the last
      compliant calibration verification shall be performed.  Information
      regarding the CCV shall be reported on Form IIA-IN.

12.3  Contract Required Quantitation Limit  (CRQL)  Check Standard  (CRI)

12.3.1   To verify linearity near the CRQL,  the Contractor shall analyze a CRI
       •  at the beginning and end of each sample analysis run, but  not before
         the ICV.  In addition, the Contractor shall analyze and report the
         results for the CRI at a frequency of not less than once per 20
         analytical samples1 per analysis run.  The CRI analysis shall be run
         immediately followed by the CCV and Continuing Calibration Blank
          (CCB) analyses.  The CRI shall be prepared by spiking an aliquot of
         reagent water with mercury at the CRQL.  The CRI shall be  taken
         through the same process used to digest and analyze the associated
         samples.

12.3.2   CRI and percent recovery results shall be reported on Form IIB-IN.
         If the percent recovery falls outside the control limits of 70-130%,
         the CRI shall be re-analyzed immediately.  If the result of the re-
         analysis falls within the control limits, no further corrective
         action is required.  If the result of the re-analysis does not fall
         within the control limits, the analysis shall be terminated, the
         problem corrected, the instrument recalibrated, or the CRI and  ,
         associated samples redigested and analyzed.

12.4  Blank Analyses

      There are two different types of blanks required by this method.   The
      calibration blank is used in establishing the analytical curve while the
      preparation blank is used to monitor for possible contamination.

12.4.1   Initial and Continuing Calibration Blank (ICB/CCB)

         The ICB and CCB are prepared with acids and reagent water  and carried
         through the entire preparation and analysis procedure.  If the
         absolute value of the calibration blank  (ICB/CCB) result exceeds the
         CRQL  (see Exhibit C), the analysis shall be terminated, the problem
         corrected, the instrument recalibrated, the calibration verified, and
         re-analysis of the preceding 10 analytical samples or all  analytical
     1As defined in Exhibit G,  CRI is an analytical sample.

                                  D-19/Mercury
ILM05.2

-------
Exhibit D (Mercury) — Section 12
Quality Control  (Con't)

         samples analyzed since the last compliant calibration blank shall be
         performed.

12.4.2   Preparation Blank  (PB)

12.4.2.1    The  PB shall contain all the reagents and in the same volumes as
            used in processing the samples.   The  PB  shall  be carried through
            the  complete procedure and contain  the same acid concentration in
            the  final  solution as the sample  solution used for analysis.

12.4.2.2    At least one PB,  consisting of reagent water processed through
            each sample preparation and analysis  procedure (see  Section  10),
            shall  be prepared and analyzed with every Sample Delivery  Group
             (SDG), or  with  each batch * of  samples digested, whichever  is more
            frequent.

12.4.2.3    The  first  batch of samples in an  SDG  is  to be  assigned to
            Preparation Blank one, the second batch  to  Preparation Blank two,
            etc.  (see  Form  III-IN).  Each  Sample  Data Package  shall contain
            the  results of  all PB analyses associated with the  samples in that
            SDG.

12.4.2.4    The  PB is  to be reported for each SDG and used in all analyses to
            ascertain  whether sample concentrations  reflect contamination in
            the  following manner:

12.4.2.4.1     If  the  absolute value of the concentration  of the blank is less
                than  or equal to  the  CRQL  (see Exhibit  C),  no further  action  is
                required.

12.4.2.4.2     If  the  analyte concentration in  the blank  is above the  CRQL,
                the lowest  concentration of the  analyte  in the  associated
                samples shall be  greater than  or equal  to 10 times  the  blank
                concentration.  Otherwise,  all samples  associated with  that
                blank,  with the analyte  concentration less than 10  times the
                blank concentration  and  above  the CRQL,  shall be redigested and
                re-analyzed with  appropriate  new Quality Control (QC).   The
                only exception to this  shall  be an identified field blank.  The
                sample concentration  is  not to be corrected for the  blank
                value.

12.4.2.4.3      If  the  concentration  of the blank is  below the  negative CRQL,
                then all samples  reported  below 10 times the CRQL and
                associated with the  blank  shall be redigested and re-analyzed
                with appropriate  new QC.

12.4.2.4.4      The values  for the PB shall be reported  on Form III-IN.

12.5  Spike Sample Analysis

12.5.1   The  spike sample  analysis is designed  to provide information  about
          the  effect  of the sample matrix  on  the digestion and/or measurement
          methodology.   The spike is  added before the digestion  (i.e.,  prior  to
          the  addition of other reagents).   At least one spike sample  analysis
          (matrix spike)  shall be performed on each group of samples  of a
          similar matrix type  (i.e.,  water,  soil)  or for each SDG.  3  The sample
       2A group of samples prepared  at  the  same  time.

       3USEPA may require additional spike  sample  analyses,  upon  USEPA Regional
 CLP Project Officer (CLP PO)  request.

 ILM05.2                          D-20/Mercury

-------
                                              Exhibit D  (Mercury) — Section 12
                                                         Quality  Control  (Con't)

         and its associated spike sample shall initially be run  at the same
         dilution.

12.5.2   If the spike analysis is performed on the same  sample that is chosen
         for the duplicate sample analysis, spike calculations shall be
         performed using the results of the sample designated as the "original
         sample"  (see Section 12.6).  The average of the duplicate results
         cannot be used for the purpose of determining percent recovery.
         Samples identified as field blanks and Performance Evaluation  (PE)
         samples shall not be used for spiked sample analysis.   USEPA may
         require that a specific sample be used for the  spike sample analysis.

12.5.3   The analyte spike shall be added at 1 ug/L (water) or 0.5 mg/kg
         (soil).  Adjustment shall be made to maintain these spiking levels
         when the weight of sample taken deviates by more  than 10% of these
         values.

12.5.4   If the spike recovery is not at or within the limits of 75-125%, the
         data of all samples received and associated with  that spike sample
         and determined by the same analytical method shall be flagged with
         the letter "N" on Forms IA-IN and VA-IN.  An exception  to this rule
         is granted when the sample concentration exceeds  the spike added
         concentration by a factor of four or more.  In  such an  event, the
         data shall be reported unflagged even if the percent recovery does
         not meet the 75-125% recovery criteria.

12.5.5   In the instance where there is more than one spike sample per matrix,
         per method, per SDG, and one spike sample recovery is not within
         contract criteria, flag all the samples of the  same matrix and method
         in the SDG.  Individual component percent recoveries  (%R) are
         calculated as follows:

         EQ. 4  Spike Percent Recovery

                           % Recovery =  SSR " SR x 100
                                           SA
         WHERE,    SSR  =  Spiked Sample Result

                   SR  =  Sample Result

                   SA  =  Spike Added

12.5.6   When sample concentration  is less than the Method Detection Limit
          (MDL), use  SR = 0 only for purposes of calculating percent recovery.
         The Spike Sample Results  (SSRs), Sample Results  (SRs), Spike Added
          (SA), and percent recovery (positive or negative) shall be reported
         on Form VA-IN.

12.5.7   The units used  for reporting SSRs will be identical to those used for
         reporting sample results on Form IA-IN.

12.6  Duplicate Sample Analysis

12.6.1   One duplicate sample  shall be analyzed from each group of samples of
         a similar matrix type (i.e., water, soil) or  for each SDG. 4
         Duplicates  cannot be  averaged for reporting on Form IA-IN.  The
     4USEPA may require additional duplicate sample analyses,  upon USEPA
Regional CLP PO request.

                                  D-21/Mercury                           ILM05.2

-------
Exhibit D  (Mercury) — Section 12
Quality Control  (Con't)

         'sample  and its associated duplicate  sample shall initially be  run at
         the  same  dilution.

12.6.2   Duplicate sample analyses are  required for percent solids.   Samples
         identified as field  blanks  and PE  samples shall not  be  used  for
         duplicate sample analysis.   USEPA  may require  that a specific  sample
         be used for  duplicate sample analysis.   The  Relative Percent
         Difference  (RPD) is  calculated as  follows:

         EQ.  5   Duplicate Sample  Relative Percent Difference

                              RPD =  'S " DI x 100
                                     (S+D)/2
          WHERE,      RPD  =  Relative Percent Difference

                       S  =  Sample Result (original)

                       D  =  Duplicate Result

 1=2.6.3    The  results  of  the  duplicate sample analyses  shall  be  reported on
          Form VI-IN.   A  control limit of 20% for RPD  shall be used for
          original and duplicate sample values greater than or equal to five
          times the CRQL  (see Exhibit C).  A control limit of the CRQL value
          shall be entered in the "Control Limit" column  on Form VI-IN if
          either the sample or duplicate value is less than five times the
          CRQL.  If the sample and duplicate values are greater  than or equal
          to five times the CRQL, or if the sample and duplicate values are
          less than the CRQL, the "Control Limit" field is left  empty.

 12.6.4    If one result is above five times the CRQL level and the other is
          below, use the  CRQL criteria to determine if the duplicate analysis  •
          is in control.   If both sample and duplicate values are less than the
          MDL, the RPD is not calculated on Form VI-IN.  For  solid sample or
          solid duplicate results less than five times the CRQL, enter the
          value of the CRQL,  corrected for sample weight and  percent solids
          (i.e., original, not duplicate sample weight and percent solids), in
          the "Control Limit" column.  If the duplicate sample results are
          outside the control limits, flag all the data for samples received
          associated with that duplicate sample with an "*" on Forms IA-IN and
          VI-IN.  In the instance where there is more  than one duplicate sample
          per SDG, if one duplicate result is not within contract criteria,
          flag all samples of the same matrix and method in the  SDG.  The
          percent difference data will be used by USEPA to evaluate the long-
          term precision of the method.  Specific control limits for each
          element will be added to Form VI-IN at a later date based on these
          precision results.

 12.7  Laboratory Control Sample (LCS) Analysis

 12.7.1    A solid LCS  (LCSS)  shall be analyzed using the same sample
          preparations, analytical methods, and Quality Assurance  (QA)/QC
          procedures employed for the EPA samples received.

 12.7.2    The USEPA provided LCSS shall be prepared and analyzed using the
          procedures applied to the solid samples received (exception: percent
          solids determination not required).  If the USEPA LCSS is
          unavailable, other USEPA QC Check samples or other certified
          materials may be used.  In  such a case, control limits for the LCSS
          must be documented and provided.  One LCSS shall be prepared and
 ILM05.2                           D-22/Mercury

-------
                                              Exhibit D (Mercury)  — Section 12
                                                        Quality Control  (Con't)

         analyzed for every group of solid samples in a SDG, or for each batch
         of samples digested, whichever is more frequent.

12.7.3   All LCSS and percent recovery results will be reported on Form VII-
         IN. If the results for the LCSS fall outside the control limits
         established by USEPA, the analyses shall be terminated, the problem
         corrected, and the samples associated with that LCSS redigested and
         re-analyzed with appropriate new QC.

12.8  Method Detection Limit  (MDL) Determination

12.8.1   Before any field samples are analyzed under this contract,  the MDLs
         shall be determined  for each digestion procedure and instrument used,
         prior to the start of contract analyses, and annually thereafter, and
         shall meet the levels specified in Exhibit C.

         An MDL study shall be performed after major instrument maintenance,
         or changes in instrumentation or instrumental conditions, to verify
         the current sensitivity of the analysis.

12.8.2   To determine the MDLs, the Contractor shall run MDL studies following
         the procedures given in 40 CFR, Part 136.  The Contractor shall
         prepare the MDL samples by each digestion procedure used and shall
         analyze these samples on each instrument used.

12.8.3   The determined concentration of the MDL shall be less than half the
         concentration of the CRQL listed in Exhibit C.

12.8.4   The results of the MDL determination studies shall be forwarded to
         the USEPA Regional CLP PO, Sample Management Office  (SMO),  and
         Quality Assurance Technical Support  (QATS).

12.8.5   The MDL results shall be reported on Form IX-IN.
                                  D-23/Mercury                          ILM05.2

-------
Exhibit D  (Mercury) — Sections 13-17
Method Performance

13.0  METHOD PERFORMANCE

      Not  applicable.

14.0  POLLUTION PREVENTION

      See  Section  1.15 in Exhibit D - Introduction to Analytical Methods.

•15.0  WASTE MANAGEMENT

      See  Section  1.16 in Exhibit D - Introduction to Analytical Methods.

16.0  REFERENCES

16.1  US Environmental Protection Agency. Methods for Chemical Analysis of
      Water and Wastes. Method 245.1. 1974.

16.2  US Environmental Protection Agency. Methods for Chemical Analysis of
      Water and Wastes. Method 245.2. 1974.

16.3  US Environmental Protection Agency. Methods for Chemical Analysis of
      Water and Wastes. Method 245.5. 1974.

16.4  US Government  Printing Office. 40 Code of Federal Regulations, Part 136,
      Section  1, Appendix B.

17.0  TABLES/DIAGRAMS/FLOWCHARTS

      Not  applicable.
 ILM05.2                          D-24/Mercury

-------
  EXHIBIT D - PART D

  ANALYTICAL METHODS
         FOR
TOTAL CYANIDE ANALYSIS
      D-1/Cyanide                           ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                           D-2/Cyanide

-------
           Exhibit D - Analytical Methods for Total  Cyanide Analysis

                               Table of Contents

Section                                                                     Page


1.0   SCOPE AND APPLICATION 	   5

2.0   SUMMARY OF METHOD	5
      2.1   Waters  and Soils   	5

3.0   DEFINITIONS  	   5

4.0   INTERFERENCES  	   6
      4.1   Sulfides   	6
      4.2   Surfactants 	   6
      4.3   Oxidizing Agents   	  6

5.0   SAFETY   	6

6.0   EQUIPMENT AND SUPPLIES   	  7
      6.1   Conventional Distillation of Water and Soils    	  7
      6.2   Midi  Distillation of Water and Soils   	7

7.0   REAGENTS AND  STANDARDS   	  8
      7.1   Reagents   	8
      7.2   Standards	9

8.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE   	  11
      8.1   Sample  Collection and Preservation   	  11
      8.2   Procedure  for Sample Storage   	  11
      8.3   Contract Required Holding Time   	  11

9.0   CALIBRATION AND STANDARDIZATION 	  12
      9.1   Instrument Operating Parameters  	  12
      9.2   General  Procedure  	  12
      9.3   Spectrophotometric Instrument Calibration Procedure  	  12
      9.4   Initial  Calibration Verification  (ICV)   	  12
      9.5   Continuing Calibration Verification  (CCV )  	  13
      9.6   Initial  and Continuing Calibration Blank (ICB/CCB)    	  13

10.0  PROCEDURE	14
      10.1  Sample  Preparation   	  14
      10.2  Water  and  Soil Preparation of Standards and Samples	15
      10.3  Sample  Analysis	19

11.0  DATA ANALYSIS AND CALCULATIONS   	20
      11.1  Water/Aqueous Sample Calculation   	  20
      ,11.2  Soil  Sample Calculatio n	20
      11.3  Calculations for Midi Distillation of Waters and Soils    ....  21
      11.4  Adjusted Method Detection Limit  (MDL)/Adjusted Contract
            Required Quantitation Limit  (CRQL) Calculation    	  22

12.0  QUALITY CONTROL  (QC)   	24
      12.1  Initial  Calibration Verification  (ICV)   	  24
      12.2  Continuing Calibration Verification  (CCV )  	  24
      12.3  Contract Required Quantitation Limit  (CRQL) Check
            Standard (CRI)   	  24
      12.4  Blank  Analyses   	24
      12.5  Spike  Sample Analysi s	25
      12.6  Duplicate  Sample Analysi s 	  27
      12.7  Laboratory Control Sample (LCS) Analysis   	  28
      12.8  Method Detection Limit  (MDL) Determination   	  28
                                   D-3/Cyanide                           ILM05.2

-------
           Exhibit  D -  Analytical  Methods  for Total Cyanide Analysis




                           Table of Contents  (Con't)




Section                                                                    Page




13.0  METHOD PERFORMANCE   	29




14.0  POLLUTION PREVENTION   	  29




15.0  WASTE MANAGEMENT   	  29




16.0  REFERENCES   	29




17.0  TABLES/DIAGRAMS/FLOWCHARTS   	  29
 ILM05.2                           D-4/Cyanide

-------
1.0
                                            Exhibit D (Cyanide)  — Sections 1-3
                                                          Scope and Application
SCOPE AND APPLICATION
      The analytical method that follows is designed to analyze various water
      types, sediment, sludge, and soil samples taken from hazardous waste
      sites, for total cyanide.

      This analytical method includes the use of acid and heat to remove
      cyanide from the sample.

2.0   SUMMARY OF METHOD

2.1   Waters and Soils

2.1.1    The cyanide as hydrocyanic acid (HCN) is released from cyanide
         complexes by means of a reflux-distillation and absorbed in a
         scrubber containing sodium hydroxide solution.  The cyanide ion in
         the absorbing solution is then determined colorimetrically.

2.1.2    In the colorimetric measurement, the cyanide is converted to cyanogen
         chloride (CNC1), by reaction with chloramine-T at a pH less than 8
         without hydrolyzing to the cyanate.  After the reaction is complete,
         color is formed on the addition of pyridine-barbituric acid reagent.
         The absorbance is read between 570 and 580 nanometers  (nm).  To
         obtain colors of comparable intensity, it is essential to have the
         same salt content in both the sample and the standards.

3.0   DEFINITIONS

      See Exhibit G for a complete list of definitions.
                                  D-5/Cyanide
                                                                  ILM05.2

-------
Exhibit D (Cyanide) — Sections 4 & 5
Interferences

4.0   INTERFERENCES

      Interferences are eliminated or reduced by using the distillation
      procedure.

4.1   Sulfides

      Sulfides  adversely affect the colorimetric procedure.  The sample should
      be tested in the field  for the presence of sulfides as described in
      Section 8.1.1.

4.2   Surfactants

      The presence of surfactants may cause the sample to foam during
      refluxing.   If this  occurs, the addition of an agent such as  Dow Corning
      544 antifoam agent will prevent the  foam from collecting in the
      condenser.

4.3   Oxidizing Agents

      Oxidizing agents such as chlorine  decompose most of the cyanides.  The
      sample  should be tested in the field for the presence of oxidizing
      agents  as described  in  Section 8.1.1.

5.0   SAFETY

      See Section  1.14 in  Exhibit D - Introduction to Analytical Methods.
 ILM05.2                           D-6/Cyanide

-------
                                               Exhibit D (Cyanide) — Section 6
                                                         Equipment and Supplies
6.0   EQUIPMENT AND SUPPLIES
      Brand names, suppliers, and part numbers are for illustrative purposes
      only.  No endorsement is implied.  Equivalent performance may be
      achieved using equipment and supplies other than those specified here,
      however, a demonstration of equivalent performance meeting the
      requirements of this Statement of Work (SOW) is the responsibility of
      the Contractor.  The Contractor shall document any use of alternate
      equipment or supplies in the Sample Delivery Group  (SDG) Narrative.

6.1   Conventional Distillation of Water and Soils

6.1.1    Reflux distillation apparatus.  The boiling flask should be of 1
         Liter  (L) size with an inlet tube and provision for condenser.  The
         gas absorber may be a Fisher-Milligan scrubber.

6.1.2    Spectrophotometer suitable for measurements between 570 and 580
         nanometers  (nm) with a 1.0 centimeter (cm) cell or larger  (for manual
         spectrophotometric method).

6.1.3    Automated analyzer instrumentation  (for automated spectrophotometric
         method) including:

6.1.3.1     Sampler

6.1.3.2     Pump

6.1.3.3     Cyanide manifold

6.1.3.4     Colorimeter with 15 millimeters  (mm)  flowcells and 580 nm filters

6.1.3.5     Recorder

6.1.3.6     Data system  (optional)

6.1.3.7     Glass or plastic tubes for the sampler

6.2   Midi Distillation of Water and Soils

6.2.1    Midi reflux distillation apparatus

6.2.2    Heating block - Capable of maintaining 125°C  (ą5°C).

6.2.3    Auto analyzer system with accessories:

6.2.3.1     Sampler

6.2.3.2     Pump

6.2.3.3     Cyanide cartridge

6.2.3.4     Colorimeter with 50 mm flowcells and 580 nm filter

6.2.3.5     Chart recorder or data system

6.2.4    Assorted volumetric glassware, pipets, and micropipets
                                  D-7/Cyanide                           ILM05.2

-------
Exhibit D (Cyanide) - Section 7
Reagents and Standards

7.0   REAGENTS AND STANDARDS

7.1   Reagents

7.1.1    Reagent water - The purity of this water must be equivalent to ASTM
         Type II water  (ASTM D1193-77).  Use this preparation for all
         reagents, standards, and  dilutions of  solutions.

7.1.2    Conventional Distillation and Preparation Reagents of Water and Soils

7.1.2.1     Sodium hydroxide solution, 1.25N -  Dissolve 50 grams  (g) of NaOH
            in reagent water, and  dilute to 1 Liter  (L) with reagent water.
             (Same  Distillation  and Preparation  Reagent  for Midi  Distillation
            of Water  and Soils.)

7.1.2.2     Cadmium carbonate - Powdered

7.1.2.3     Ascorbic  acid  - Crystals

7.1.2.4     Sulfuric  acid  - Concentrated

7.1.2.5     Hydrochloric acid  (HC1) -  Concentrated  (specific gravity 1.19).

7.1.2.6     Magnesium chloride  solution -  Weigh 510  g of MgCl2  • 6H20  into a
             1000 milliliter  (mL)  flask, dissolve,  and dilute to  1  L with
             reagent water.  (Same  Distillation  and Preparation Reagent for
             Midi Distillation  of Water and Soils.)

7.1.3    Midi Distillation and  Preparation Reagents  of  Water and Soils

7.1.3.1      Sodium hydroxide absorbing solution and sample wash  solution,
             0.25N  -  Dissolve  10.0  g NaOH  in  reagent water  and dilute  to 1  L.

7.1.3.2      Sulfuric  acid, 50%  (v/v)  - Carefully add a  portion of  concentrated
             H2S04 to an equal portion of reagent water.

7.1.4    Manual  Spectrophotometric Reagents  for Water and  Soils

7.1.4.1      Acetate Buffer - Dissolve 410  g  of  NaC  2H302 •  3H20 in 500  mL of
             reagent  water.  Add sufficient glacial acetic  acid  to adjust  pH to
             4.5  (approximately 500 mL).

7.1.4.2      Chloramine-T solution  - Dissolve  1.0 g of white, water soluble
             chloramine-T in  100 mL of reagent water and refrigerate until
             ready to use.   Prepare fresh  weekly.

7.1.4.3      Color  Reagent

7.1.4.3.1       Pyridine-barbituric acid reagent - Place 15 g of  barbituric
                acid in a  250 mL volumetric flask and add just enough reagent
                water to wash the  sides of the flask and wet the barbituric
                acid.   Add 75 mL of pyridine and mix.  Add 15 mL of HC1
                (specific  gravity  1.19), mix,  and cool to room temperature.
                Dilute to  250 mL with reagent water and mix.   This reagent is
                stable for approximately six months if stored in a cool, dark
                place.

7.1.5     Semi-Automated Spectrophotometric Reagents for Conventional  and  Midi
          Distillation of Water and Soils

7.1.5.1      Chloramine-T solution -  Dissolve 0.40 g of  chloramine-T in reagent
             water and dilute to 100 mL.  Prepare fresh daily.


 ILM05.2                            D-8/Cyanide

-------
                                                Exhibit D (Cyanide) - Section 7
                                                 Reagents and Standards  (Con't)

7.1.5.2     Acetate Buffer - Dissolve 410 g of NaC 2H3O2 • 3H20 in 500 mL of
            reagent water.  Add sufficient glacial acetic acid to  adjust pH to
            4.5  (approximately 500 mL).

7.1.5.3     Pyridine-barbituric acid solution - Transfer 15 g of barbituric
            acid into a 1 liter volumetric flask.  Add about  100 mL of  reagent
            water and swirl the flask.  Add 75 mL of pyridine and  mix.  Add 15
            mL of concentrated HC1 and mix.  Dilute to about  900 mL with
            reagent water and mix until the barbituric acid is dissolved.
            Dilute to 1 L with reagent water.  Store at 4°C  (ą2°C).

7.1.5.4     Sampler wash - Dissolve 10 g of NaOH in reagent water  and dilute
            to 1 L.   (For conventional distillation of water  and soils  only.)

7.2   Standards

7.2.1    Introduction

         The Contractor must provide all standards to be used with this
         contract.  These standards may be used only after they have been
         certified according to the procedure in Exhibit E, Section 8.0.  The
         Contractor must be able to verify that the standards are  certified.
         Manufacturer's certificates of analysis must be retained  by the
         Contractor and presented upon request.

7.2.2    Stock Standard Solutions

7.2.2.1     Stock Standard Reagents for Water and Soils

7.2.2.1.1      Stock cyanide solution - Dissolve 2.51 g of KCN and 2 g KOH in
               1 L of reagent water.  Standardize with 0.0192N AgN03.

7.2.2.1.2      Standard cyanide solution, intermediate - Dilute 50.0 mL of
               stock  (1 mL = 1 milligram  (mg) CN) to 1000 mL  with  reagent
               water.

7.2.2.1.3      Standard cyanide solution - Prepare fresh daily by  diluting 100
               mL of  intermediate cyanide solution to 1000 mL with reagent
               water  and  store in a glass stoppered bottle.   1 mL  =  5.0
               micrograms  (pg) CN  [5.0 milligrams per Liter  (mg/L)].

7.2.2.1.4      Sodium hydroxide solution, 0.25N - Dissolve 10 g of NaOH in
               reagent water and dilute to 1 L.

7.2.2.2     Stock Standard Reagents for Midi Distillation of Water and Soils

7.2.2.2.1      Stock cyanide solution, 1000 mg/L CN - Dissolve 2.51 g of KCN
               and 2.0 g  KOH in reagent water and dilute 1 L.  Standardize
               with 0.0192N AgN03.

7.2.2.2.2      Intermediate cyanide standard solution, 10 mg/L CN  - Dilute 1.0
               mL of  stock cyanide solution  (see Section 7.2.2.2.1) plus 20 mL
               of 1.25N NaOH solution  (see Section 7.1.2.1) to 100 mL with
               reagent water.  Prepare this solution at time  of analysis.

7.2.2.2.3      Sodium hydroxide solution, 0.IN - Dissolve 4 g of NaOH in
               reagent water and dilute to 1 L.
                                  D-9/Cyanide                            ILM05.2

-------
Exhibit D (Cyanide) - Section 7
Reagents and Standards  (Con't)

7.2.3    Secondary Dilution Standards

7.2.3.1     Secondary Dilution Standards

            Prepare  secondary dilution standard solutions by diluting the
            appropriate volumes of  stock standards with  0.25N NaOH.  The final
            concentration  of  NaOH'in all standards should be 0.25N.

7.2.4    Working Standards

7.2.4.1     Method Detection  Limit  (MDL) Solution

7.2.4.1.1      The MDL  solution shall be at a concentration of 3 to 5 times
               the expected MDL.

7.2.4.2     Contract Required Quantitation Limit  (CRQL)  Check Standard  (CRI)

7.2.4.2.1      The concentration of the CRI for cyanide  shall be at the CRQL.
               Information regarding  the CRI  shall be  reported on  Form  IIB-IN.
 ILM05.2                          D-10/Cyanide

-------
                                               Exhibit D (Cyanide)  — Section 8
                                   Sample Collection,  Preservation,  and Storage

8.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE

8.1   Sample Collection and Preservation

8.1.1    Water Sample Preservation

         Collection of  total cyanide must be in polyethylene or glass
         containers.  The sample must be tested for sulfides and oxidizing
         agents, and preserved by the sampler immediately upon sample
         collection.  Place a drop of the sample on lead acetate test paper
         (which has been pre-moistened with pH 4 acetate buffer solution) to
         detect the presence of sulfides.  If sulfides are present  (test strip
         turns black),  the sample volume required for the cyanide
         determination  should be increased by 25 milliliters (mL).  The total
         volume of sample should then be treated with powdered cadmium
         carbonate or lead carbonate.  Yellow cadmium sulfide precipitates if
         the sample contains sulfide.  Repeat this operation until a drop of
         the treated sample solution does not darken the lead acetate test
         paper.  Filter the solution through a dry filter paper into a dry
         beaker, and from the filtrate measure the sample to be used for
         analysis.  Avoid a large excess of cadmium carbonate and a long
         contact time in order to minimize a loss by complexation or occlusion
         of cyanide on  the precipitated material.  If no sulfides are present,
         test for the presence of oxidizing agents by placing a drop of the
         sample on a strip of potassium iodide - starch test paper  (KI -
         starch paper); a blue color indicates the need for treatment.  Add
         ascorbic acid, a few crystals at a time, until a drop of sample
         produces no color on the indicator paper.  Then add an additional 0.6
         gram  (g) of ascorbic acid for each liter of sample volume.  Preserve
         the sample with NaOH to pH greater than 12 and maintain at 4°C  (ą2°C)
         until distillation,

8.1.2    Soil/Sediment  Sample Preservation

         Samples shall  be kept at 4°C  (ą2°C) from the time of collection until
         distillation.

8.2   Procedure for Sample Storage

8.2.1    The samples must be protected from light and refrigerated at 4°C
         (ą2°C) from the time of receipt until 60 days after delivery of a
         complete, reconciled data package to the USEPA.  After 60 days the
         samples may be disposed of in a manner that complies with all
         applicable regulations.

8.2.2    The samples must be stored in an atmosphere demonstrated to be free
         of all potential contaminants.

8.2.3    Samples, sample distillates, and standards must be stored separately.

8.3   Contract Required Holding Time

         The maximum sample holding time for cyanide is 12 days from Validated
         Time of Sample Receipt  (VTSR).
                                  D-11/Cyanide                          ILM05.2

-------
Exhibit D (Cyanide) — Section 9
Calibration and Standardization

9.0   CALIBRATION AND STANDARDIZATION

9.1   Instrument Operating Parameters

      Because of the difference between various makes and models of
      satisfactory instruments, no detailed operating instructions can be
      provided.  The analyst should follow the instructions provided by the
      manufacturer of the particular instrument.  It is the responsibility of
      the analyst to verify that the instrument configuration and operating
      conditions used satisfy the analytical requirements and to maintain
      Quality Control  (QC) data confirming instrument performance and
      analytical results.

9.2   General Procedure

      The following general procedure applies to most semi-automated
      colorimeters.  Set up the manifold and complete system per
      manufacturer's instructions.  Allow the colorimeter and recorder to warm
      up for at least 30 minutes prior to use.  Establish a steady reagent
      baseline, feeding reagent water through the sample line and appropriate
      reagents  (see Section 7.1.5) through reagent lines.  Adjust the baseline
      using the appropriate control on the colorimeter.  Prepare a standard
      curve by plotting absorbance of standard vs. cyanide concentrations  [per
      250 milliliter  (mL)].

9.3   Spectrophotometric  Instrument Calibration Procedure

9.3.1    Instruments shall be  calibrated daily or once every 24 hours, and
         each  time the  instrument  is  set up.  The instrument standardization
         date  and  time  shall be included in the raw  data.

9.3.2    The date  and time of  preparation and analysis shall be given in the
         raw data.

9.3.3    Calibration standards  shall be prepared fresh daily or each time an
         analysis  is to be made and discarded after  use.  Prepare a blank  and
         at least  three calibration standards in graduated amounts in the
         appropriate range.  One  of the calibration  standards  shall be at  the
         Contract  Required Quantitation Limit  (CRQL).  The acceptance criteria
         for the initial  calibration  curve  is a correlation coefficient
         greater than  or  equal to  0.995.

9.3.4    Any changes or corrections to the  analytical system shall be followed
         by recalibration.

9.4    Initial  Calibration Verification  (ICV)

9.4.1    Immediately after each cyanide system has been  calibrated, the
         accuracy  of  the  initial  calibration  shall be verified and documented
         for cyanide by the  analysis  of the ICV Solution at the wavelength
         used  for  analysis.

9.4.2    Only  if the ICV  Solution  is  not available from  USEPA,  or where a
         certified solution  of the analyte  is not available from any  source,
         analyses  shall be  conducted  on an  independent  standard at a
         concentration other than that used for instrument calibration, but
         within the  calibration range.  An  independent  standard is defined as
         a standard  composed of the analytes  from a  different  source  than
         those used  in the  standards  for  the  instrument  calibration.
 ILM05.2                          D-12/Cyanide

-------
                                               Exhibit D (Cyanide)  — Section 9
                                        Calibration and Standardization (Con't)

9.4.3    The ICV shall be distilled.  This means that an ICV must be distilled
         with each batch of samples analyzed and that the samples distilled
         with an ICV must be analyzed with that particular ICV.

9.4.4    The value for the ICV shall be reported on Form IIA-IN.

9.5   Continuing Calibration Verification (CCV)

9.5.1    To ensure calibration accuracy during each analysis run, one of the
         following standards is to be used for the CCV and shall be analyzed
         and reported at a frequency of 10% or every 2 hours during an
         analysis run, whichever is more frequent.  The standard shall also be
         analyzed and reported at the beginning of the run and after the last
         analytical sample.  The analyte concentration in the CCV standard
         shall be different than the concentration used for the ICV and shall
         be one of the following solutions at or near the mid-range level of
         the calibration curve:

         =*=    USEPA  Solutions

         :i:    NIST Standards

         ™=    A Contractor-prepared standard  solution

         The same CCV standard shall be used throughout the analysis runs for
         a Sample Delivery Group (SDG) of samples received.

9.5.2    Each CCV analyzed shall reflect the conditions of analysis of all
         associated analytical samples  (the preceding 10 analytical samples or
         the preceding analytical samples up to the previous CCV).  The
         duration of analysis, rinses, and other related operations that may
         affect the CCV measured result may not be applied to the CCV to a
         greater extent than the extent applied to the associated analytical
         samples.  For instance, the difference in time between a CCV analysis
         and the blank immediately following it, as well as the difference in
         time between the CCV and the analytical sample immediately preceding
         it, may not exceed the lowest difference in time between any two
         consecutive analytical samples associated with the CCV.

9.5.3    Information regarding the CCV shall be reported on Form IIA-IN.

9.6   Initial and Continuing Calibration Blank (ICB/CCB)

      A calibration blank shall be analyzed at the wavelength used for
      analysis immediately after every ICV and CCV, at a frequency of 10% or
      every 2 hours during the run, whichever is more frequent.  The blank
      shall be analyzed at the beginning of the run and after the last
      analytical sample.

      NOTE: A CCB shall be analyzed immediately after the last CCV, and the
      last CCV shall be analyzed immediately after the last analytical sample
      of the run.  The results for the calibration blanks shall be reported on
      Form III-IN.
                                  D-13/Cyanide                          ILM05.2

-------
Exhibit D (Cyanide) — Section 10
Procedure

10.0  PROCEDURE

10.1  Sample Preparation

10.1.1   If insufficient  sample amount  (less than  90%,  of  the required amount)
         is received to perform the analyses,  the  Contractor shall contact
         Sample Management  Office  (SMO)  to  inform  them of  the problem.   SMO
         will contact  the Region for instructions.   The Region  will either
         require that  no  sample analyses be performed or will require that a
         reduced volume be  used for the  sample analysis.  No other changes in
         the analyses  will  be permitted.  The  Contractor shall  document  the
         Region's  decision  in the  Sample Delivery  Group (SDG) Narrative.

10.1.2   If multiphase samples  (e.g., two-phase  liquid sample,  oily
         sludge/sandy  soil  sample) are  received by the Contractor, the
         Contractor shall contact  SMO to apprise them of the type of  sample
         received.  SMO will contact the Region.  If all phases of the sample
         are amenable  to  analysis, the  Region  may  require  the Contractor to do
         any of the following:

         :i:    Mix the sample  and  analyze  an  aliquot from  the homogenized
               sample.

         =*=    Separate  the phases of  the  sample,  and analyze one  or  more of
               the phases separately.   SMO will provide EPA sample numbers for
               the additional  phases,  if required.

         1X3    Do  not  analyze  the  sample.

10.1.2.1    If all of  the phases are not amenable  to analysis  (i.e.,  outside
            scope), the  Region  may require  the Contractor  to do any of the
            following:

            ^     Separate the phases  and analyze the phase(s)  that  is   (are)
                   amenable to analysis.  SMO will provide EPA sample  numbers
                   for the additional phases,  if required.

            :*:     Do  not analyze the sample.

10.1.2.2    No other changes in  the  analyses  will  be permitted. The
            Contractor shall document  the  Region's decision in  the SDG
            Narrative.

10.1.3   Soil samples  are not  dried  prior  to  analysis.  A  separate percent
         solids  determination  must be  made in accordance with the procedure  in
         Exhibit D -  Introduction  to Analytical Methods, Section 1.6.

10.1.4   Before preparation is  initiated for  an aqueous sample, the Contractor
         shall test for the presence of sulfides and oxidizing  agents (e.g.,
         residual  chlorine).  The  test for sulfides shall  be  performed by
         placing a drop of the sample  on a strip of lead acetate paper  (which
         has  been  pre-moistened with pH 4  acetate buffer solution).   If  the
         test strip turns black,  the Contractor shall treat the total volume
         of sample with powdered cadmium carbonate or lead carbonate.  Yellow
         cadmium sulfide precipitates  when the sample contains  sulfide.   This
         operation shall be repeated until a drop of the treated sample
         solution  does not darken  the  lead acetate test paper.   The  solution
          shall be  filtered through a dry filter paper into a  dry beaker, and
         the  volume of sample  to be  used for analysis shall be  measured  from
         the  filtrate.  It is  recommended that the Contractor  avoid  a large
         excess  of cadmium carbonate and a long contact time  in order to
         minimize  a loss by complexation or occlusion of cyanide on  the


 ILM05.2                           D-14/Cyanide

-------
                                              Exhibit D (Cyanide)  — Section 10
                                                              Procedure  (Con't)

         precipitated material.  The test for oxidizing agents shall be
         performed by placing a drop of the sample on a strip of potassium
         iodide - starch test paper  (KI - starch paper).  If the test strip
         turns blue, the Contractor shall contact SMO for further instructions
         from the Region before proceeding with sample preparation and
         analysis.  The Contractor shall document the presence of sulfides or
         oxidizing agents in the SDG Narrative.

10.2  Water and Soil Preparation of Standards and Samples

10.2.1   Standards Preparation

10.2.1.1    It is not imperative that all standards be distilled in the same
            manner as the samples.  At least one standard  (mid-range) must be
            distilled and compared to similar values on the curve to ensure
            that the distillation technique is reliable.  If the distilled
            standard does not agree within ą15% of the undistilled standards,
            the operator shall find and correct the cause of the apparent
            error before proceeding.

10.2.1.2    Standards for Manual Spectrophotometric Analysis of Water and Soil
            Samples

            Prepare a minimum of three standards and a blank by pipetting
            suitable volumes of standard solution into 250 milliliter  (mL)
            volumetric flasks.

            NOTE: The concentration of one of the calibration standards shall
            be at the Contract Required Quantitation Limit  (CRQL).

            To each standard, add 50 mL of 1.25N  NaOH and dilute to 250 mL
            with reagent water.  The same method for color development  (i.e.,
            pyridine-barbituric acid or pyridine-pyrazolone) must be used for
            both the samples and standards.  Standards must bracket the
            concentration of the samples.  If dilution is required, use the
            blank solution.

10.2.1.3    Standards for Semi-Automated Spectrophotometric Analysis of Water
            and Soil Samples

            Calibration standards - Prepare a blank and at least three
            calibration standards over the range of the analysis by pipetting
            suitable volumes of standard solution into volumetric flasks.  One
            calibration standard must be at the CRQL.  Add NaOH to each
            standard to bring the concentration of NaOH to 10 grams per Liter
             (g/L).  Store at 4°C  (ą2°C).

10.2.1.4    Standards for Midi Distillation Preparation and Semi-Automated
            Spectrophotometric Analysis of Water and Soil Samples

            Prepare a minimum of three standards and a blank by pipetting
            suitable volumes of standard solution into 50 mL volumetric
            flasks.  Dilute standards to 50 mL with 0.25N NaOH.

            NOTE: One calibration standard must be at the CRQL.

10.2.2   Water Samples Preparation  (Distillation)

10.2.2.1    Preparation Method/Code  (DW1)

10.2.2.1.1     Place 500 mL of sample in the 1 liter boiling flask.  Add 50 mL
               of NaOH solution  (see Section 7.1.2.1) to the absorbing  tube


                                  D-15/Cyanide                           ILM05.2

-------
Exhibit D (Cyanide) — Section 10
Procedure (Con't) •

               and  dilute  if  necessary with  reagent water  to obtain an
               adequate  depth of  liquid  in the  absorber.   Connect  the boiling
               flask,  condenser,  absorber and trap in  the  train.

10.2.2.1.2     Start a slow stream of air entering the boiling flask by
               adjusting the  vacuum  source.  Adjust the vacuum so  that
               approximately  one  bubble  of air  per second  enters the boiling
               flask through  the  air inlet tube.

               NOTE: The bubble rate will not remain constant after the
               reagents  have  been added  and  while heat is  being applied  to  the
               flask.   It  will be necessary  to  re-adjust the air rate
               occasionally to prevent the solution in the boiling flask from
               backing up  into the air inlet tube.

10.2.2.1.3     Slowly  add  25  mL concentrated sulfuric  acid (H2 S04) (see
               Section 7.1.2.4) through  the  air inlet  tube.  Rinse the tube
               with reagent water and allow  the airflow to mix the flask
               contents  for three minutes.   Pour 20 mL of  magnesium chloride
               solution  (see  Section 7.1.2.6) into the air inlet and wash down
               with a  stream  of water.

10.2.2.1.4     Heat the  solution  to  boiling, taking care to prevent the
               solution  from  backing up  into and overflowing from  the air
               inlet tube. Reflux for one hour.  Turn off heat and continue
               the airflow for at least  15 minutes.  After cooling the boiling
               flask,  disconnect  absorber and close off the vacuum source.

10.2.2.1.5     Drain the solution from the absorber into a 250 mL  volumetric
               flask and bring up to volume  with reagent water washings  from
               the absorber tube.

               NOTE: The distillation procedure results in a  two-fold
               concentration  of  the  sample.

10.2.3   Water Samples Preparation  (Midi-Distillation)

10.2.3.1     Preparation  Method/Code  (DW2)

10.2.3.1.1     The procedure  described here  utilizes a midi distillation
               apparatus and  requires a  sample  aliquot of  50 mL or less  for
               aqueous samples.

10.2.3.1.2     Pipet 50  mL of sample, or an  aliquot diluted to 50  mL, into  the
               distillation flask along  with 2  or  3 boiling chips.

10.2.3.1.3     Add 50  mL of 0.25N NaOH  (see  Section 7.1.3.1) to the gas
                absorbing impinger.

10.2.3.1.4     Connect the boiling flask, condenser, and absorber  in the
                train.   The excess cyanide  trap  contains  0.5N NaOH.

10.2.3.1.5     Turn on the vacuum and adjust the gang  (Whitney) valves to give
                a flow of three bubbles  per  second from the impingers  in  each
                reaction vessel.

10.2.3.1.6     After  five  minutes of vacuum  flow,  inject 5 mL of 50%  (v/v)
                H2S04 (see  Section 7.1.3.2)  through the top air inlet tube of
                the distillation head into  the reaction vessel.   Allow  to mix
                for 5  minutes.
 ILM05.2                           D-16/Cyanide

-------
                                              Exhibit D (Cyanide)  — Section 10
                                                              Procedure  (Con't)

               NOTE: The acid volume must be sufficient to bring the
               sample/solution pH to below 2.0.

10.2.3.1.7     Add 2 mL of magnesium chloride solution (see Section 7.1.2.6)
               through the top air inlet tube of the distillation head into
               the reaction flask.  Excessive foaming from samples containing
               surfactants may be quelled by the addition of either another 2
               mL of magnesium chloride solution or a few drops of a
               commercially available anti-foam agent.  The Contractor shall
               document the addition of magnesium chloride solution or anti-
               foam agent in the SDG Narrative.

10.2.3.1.8     Turn on the heating block and set for 123-125°C.   Heat the
               solution to boiling, taking care to prevent solution backup by
               periodic adjustment of the vacuum flow.

10.2.3.1.9     After one and a half hours of refluxing, turn off the heat and
               continue the vacuum for an additional 15 minutes.  The flasks
               should be cool at this time.

10.2.3.1.10    After cooling,  close off the vacuum at the gang valve and
               remove the absorber.  Seal the receiving solutions and store
               them at 4°C until analyzed.  The solutions must be analyzed  for
               cyanide within the 12 day holding time specified in Section
               8.3.

10.2.4   Soil Samples Preparation

10.2.4.1    Preparation Method/Code (DS1)  (Distillation)

10.2.4.1.1     Accurately weigh a representative 1-5 gram (g)  portion of wet
               sample and transfer it to a boiling flask.  Add 500 mL of
               reagent water'.  Shake or stir the sample so that it is
               dispersed.

10.2.4.1.2     Add 50 mL of NaOH solution (see Section 7.1.2.1)  to the
               absorbing tube and dilute if necessary with re'agent water to
               obtain an adequate depth of liquid in the absorber.  Connect
               the boiling flask, condenser, absorber, and trap in the train.

10.2.4.1.3     Start a slow stream of air entering the boiling flask by
               adjusting the vacuum source.  Adjust the vacuum so that
               approximately one bubble of air per second enters the boiling
               flask through the air inlet tube.

               NOTE: The bubble rate will not remain constant after the
               reagents have been added and while heat is being applied to  the
               flask.  It will be necessary to re-adjust the air rate
               occasionally to prevent the solution in the boiling flask from
               backing up into the air inlet tube.

10.2.4.1.4     Slowly add 25 mL of concentrated H 2SO4  (see Section 7.1.2.4)
               through the air inlet tube.  Rinse the tube with reagent water
               and allow the airflow to mix the flask contents for 3 minutes.
               Pour 20 mL of magnesium chloride solution  (see Section 7.1.2.6)
               into the air inlet and wash down with a stream of water.

10.2.4.1.5     Heat the solution to boiling,  taking care to prevent the
               solution from backing up and overflowing into the air inlet
               tube.  Reflux for one hour.  Turn off heat and continue the
               airflow for at least 15 minutes.  After cooling the boiling
               flask, disconnect absorber and close off the vacuum source.


                                  D-17/Cyanide                          ILM05.2

-------
Exhibit D (Cyanide) — Section 10
Procedure (Con't)

10.2.4.1.6     Drain the solution from the absorber into a 250 mL.volumetric
               flask and bring up ,to volume  with  reagent water washings  from
               the absorber  tube.

10.2.4.2    Preparation Method/Code  (DS2)  (Midi-Distillation)

10.2.4.2.1     The procedure described here  utilizes a midi distillation
               apparatus and requires a  sample aliquot of  1 gram for solid
               materials.

10.2.4.2.2     Weigh 1.0 g of sample  (to  the nearest 0.01 g)  into the
               distillation  flask  and dilute to 50  mL with  reagent  water.   Add
               2 or 3  boiling chips.

10.2.4.2.3     Add 50  mL of  0.25N NaOH  (see  Section 7.1.3.1)  to  the gas
               absorbing impinger.

10.2.4.2.4     Connect the boiling  flask,  condenser, and absorber in the
               train.  The  excess  cyanide trap contains  0.5N  NaOH.

10.2.4.2.5     Turn on the vacuum and adjust the  gang  (Whitney)  valves to  give
               a flow  of three bubbles  per second from the  impingers in  each
               reaction vessel.

10.2.4.2.6     After five minutes of vacuum  flow, inject 5 mL of 50% (v/v)
               H2SO4 (see Section 7.1.3.2) through the top air inlet tube of
               the distillation  head into the  reaction vessel.   Allow to mix
                for 5 minutes.

               NOTE:  The  acid volume must be sufficient  to bring the
                sample/solution pH to below 2.0.

10.2.4.2.7     Add 2 mL of  magnesium chloride  solution  (see Section 7.1.2.6)
                through the top  air inlet tube  of the  distillation head into
                the reaction flask.   Excessive  foaming from samples  containing
                surfactants may be quelled by the addition of  either another 2
               mL of magnesium  chloride solution or a few drops of  a
                commercially available anti-foam agent.   The Contractor shall
                document  the addition of magnesium chloride solution or anti-
                foam agent  in the SDG Narrative.

10.2.4.2.8     Turn on the  heating block and set for  123-125°C.   Heat the
                solution  to boiling, taking care to prevent solution backup by
                periodic  adjustment of the vacuum flow.

10.2.4.2.9     After one and a half hours of refluxing,  turn  off the heat  and
                continue  the vacuum for an additional  15  minutes.  The flasks
                should be cool at this time.

10.2.4.2.10   After cooling, close off the  vacuum  at  the  gang  valve and
                remove the absorber.  Seal the receiving  solutions and store
                them at 4°C until analyzed.  The solutions must be analyzed for
                cyanide within the 12 day holding time specified in
                Section 8.3.

 10.2.5   Non-Distilled Analyses

 10.2.5.1    Preparation  Method/Code (NP1)

 10.2.5.1.1      This code shall  be used to report samples that are not
                distilled prior to analysis.
 ILM05'.2                           D-18/Cyanide

-------
                                              Exhibit D (Cyanide) — Section 10
                                                               Procedure  (Con't)

10.2.5.1.2     This Preparation Method/Code shall also be used to report the
               non-distilled Method Detection Limit  (MDL).  The  concentration
               of this MDL shall be used  to determine the appropriate
               concentration qualifier  for the  results of instrument QC
               analyses  [except the distilled Initial Calibration Verification
                (ICV)].

10.3  Sample Analysis

10.3.1   Manual Spectrophotometric Determination

10.3.1.1    Allow all standards and samples to  come to ambient room
            temperature prior to analysis.  Withdraw 50 mL or  less of  the
            solution from the flask and transfer to a 100 mL volumetric  flask.
            If less than 50 mL is taken,  dilute to 50 mL with  0.25N sodium
            hydroxide solution  (see Section 7.1.3.1).  Add 1.0 mL of acetate
            buffer  (see Section 7.1.4.1)  and mix.  The dilution  factor must be
            reported on Form XIII-IN.

10.3.1.2    Add 2 mL of chloramine-T  (see Section 7.1.4.2) and mix.  After 1
            to 2 minutes, add 5 mL of   pyridine-barbituric acid  solution  (see
            Section 7.1.4.3.1) and mix.   Dilute to mark with reagent water and
            mix again.  Allow 8 minutes for color development  then read
            absorbance between 570 and  580 nanometers  (nm) in  a  1 centimeter
             (cm) cell within 15 minutes.

10.3.2   Semi-Automated Spectrophotometric Determination of Distillates

10.3.2.1    Set up the manifold.  Pump  the reagents through the  system until a
            steady baseline is obtained.

10.3.2.2    Place calibration standards,  blanks, and control standards in the
            sampler tray, followed by distilled samples, distilled duplicates,
            distilled standards, distilled spikes, and distilled blanks.
            Allow all standards and samples to  come to ambient room
            temperature prior to analysis.

10.3.2.3    When a steady reagent baseline is-obtained and before starting the
            sampler, adjust the baseline  using  the appropriate knob on the
            colorimeter.  Aspirate a calibration standard and  adjust the
            colorimeter until the desired signal is obtained.  Establish the
            baseline and proceed to analyze calibration standards, blanks,
            control standards, distilled  samples, and distilled  Quality
            Control  (QC) samples.
                                  D-19/Cyanide                           ILM05.2

-------
Exhibit D (Cyanide) — Section 11
Data Analysis and Calculations

11.0  DATA ANALYSIS AND CALCULATIONS

11.1  Water/Aqueous Sample Calculation

11.1.1   For semi-automated colorimetric determination  (Non-Midi-
         Distillation) , measure  the instrument response of  the  calibration
         standards and calculate a linear  regression equation.  Apply  the
         equation  to  the  samples and Quality Control  (QC) samples  to determine
         the cyanide  concentration in the  distillates.  To  determine the
         concentration of cyanide in the original  sample, MULTIPLY THE RESULTS
         BY ONE-HALF  (since the  original volume was 500 milliliter (mL) and
         the distillate volume was 250 mL).  Also  correct for,  and report on
         Form XIII-IN, any dilutions which were made before or  after
         distillation.

11.1.2   For manual colorimetric determination, calculate the cyanide,  in
         micrograms per Liter (ug/L), in the original  sample as follows:

         EQ. 1   Aqueous Sample Concentration  (Manual)

                 „„ „      ^  _.       /T.    A x 1000 mL/L   50 mL
                 CN Concentration (ug/L) =  	 x 	
                                                B            C
          WHERE,

           A           =   pg CN read from standard curve (per 250 mL)

           B           =   mL of original sample for distillation (see Section
                           10.2.2.1.1)

           C           =   mL taken for colorimetric analysis (see Section
                           10.3.1.1)

           50 mL       =   volume of  original sample aliquot (see Section
                           10.3.1.1)

           1000 mL/L   =   conversion mL to L

          The minimum value that can  be substituted for A is the Method
          Detection Limit (MDL)  value adjusted for volume.

 11.2  Soil Sample Calculation

 11.2.1   A separate determination of percent solids must be performed (see
          Exhibit D - Introduction to Analytical Methods, Section 1.6).

 11.2.2   The concentration of cyanide in the sample is determined as follows:

 11.2.2.1    Manual Spectrophotometric

             EQ. 2  Soil Sample Concentration  (Manual)
                     CN Concentration (mg/kg)  =
                                                      50 mL
                                                  A x 	
                                                    % solids
                                                       100
 ILM05.2                           D-20/Cyanide

-------
            WHERE,

              A

              B
              50 mL
              % solids   =
                                              Exhibit D (Cyanide) — Section 11
                                         Data Analysis and Calculations (Can't)
 pg CN  read  from  standard  curve  (per 250  mL).

 mL of  distillate taken  for  colorimetric
 determination  (see  Section  10.3.1.1).

 wet  weight  of  original  sample in  g (see  Section
 10.2.4.1.1).

 standard volume  taken for colorimetric
 determination  (see  Section  10.3.1.1)

 percent solids  (see Exhibit  D - Introduction  to
 Analytical  Methods, Section 1.6).
11.2.2.2    Semi-Automated Spectrophotometric for Non-Midi-Distillates

            If the semi-automated method is used, measure the peak heights of
            the calibration standards  (visually or using a data  system) and
            calculate a linear regression equation.  Apply the equation to the
            samples and QC audits to determine the cyanide concentration  in
            the distillates.

            EQ. 3  Soil Sample Concentration  (Semi-automated)

                                                  A x .25
                    CN Concentration (mg/kg)
                                               C x
                        % solids
                          100
            WHERE,

             A

             C


             .25


             % solids
pg/L determined from standard curve.

wet weight of original sample in g  (see Section
10.2.4.1.1) .

conversion factor for distillate final volume
(see Section 10.2.4.1.6).

percent solids (see Exhibit D - Introduction to
Analytical Methods, Section 1.6).
            The minimum value that can be substituted for A is the MDL value.

11.3  Calculations for Midi Distillation of Waters and Soils

11.3.1   Calculations for Semi-automated Colorimetric Determination

11.3.1.1    Prepare a standard curve by plotting absorbance (peak heights,
            determined visually or using a data system) of standards  (y)
            versus cyanide concentration values  (total pg CN/L)  (x).  Perform
            a linear regression analysis.

11.3.1.2    Multiply all distilled values by the standardization value to
            correct for the stock cyanide solution not being exactly  1000
            milligrams per Liter  (mg/L)  (see Section 7.2.2.2.1).

11.3.1.3    Using the regression analysis equation, calculate sample  receiving
            solution concentrations from the calibration curve.

11.3.1.4    Calculate the cyanide of aqueous samples in pg/L of original
            sample, as follows:
                                  D-21/Cyanide
                                             ILM05.2

-------
Exhibit D  (Cyanide) — Section 11
Data Analysis and Calculations (Con't)

            EQ.  4  Aqueous  Sample  Concentration  (Midi)

                                                 A x D x F
                       CN Concentration (pg/L)  =
                                                     B
            WHERE,

              A   =     ug/L  CN  of  sample  from regression  analysis

              B   =     volume of original sample  for distillation  (0.050  L)
                        (see  Section 10.2.3.1.2)

              D   =     any dilution factor necessary to bracket  sample  value
                        within standard values

              F   =     sample receiving solution  volume (0.050 L)

            The minimum value  that can be substituted for A is  the MDL  value.

 11.3.1.5    Calculate the cyanide  of solid samples in mg/kg of original
            sample,  as follows:

 11.3.1.5.1      A separate determination of percent solids must be  performed
                 (see Exhibit D - Introduction to Analytical Methods, Section
                1.6).

 11.3.1.5.2      The concentration of cyanide in the sample is determined as
                follows:

                EQ. 5  Soil Sample  Concentration  (Midi)

                                                  A x D x F
                       CN Concentration (mg/kg)  =
                                                    B x E
                WHERE,

                 A    =     pg/L CN of sample from regression analysis curve

                 B    =     wet weight of original sample  (see Section
                           10.2.4.2.2)

                 D    =     any dilution factor necessary to bracket sample
                           value within standard values

                 E         % solids/100 (see Exhibit D - Introduction to
                           Analytical Methods, Section 1.6)

                 F    =     sample receiving solution volume  (0.050 L)

                The minimum value that can be substituted for A is the MDL
                value.

 11.4  Adjusted Method Detection Limit (MDL)/Adjusted Contract Required
       Quantitation Limit  (CRQL) Calculation

       To calculate the adjusted aqueous MDL or adjusted aqueous CRQL for the
       manual colorimetric method, multiply the MDL  (pg/L) or CRQL (pg/L) by
       0.25 and substitute the result for the "A" term in Equation 1.  To
       calculate the adjusted aqueous MDL or adjusted aqueous CRQL for all
       other methods, follow the instructions in Section 11.1.1 or substitute

 ILM05.2                           D-22/Cyanide

-------
                                        Exhibit D (Cyanide) — Section 11
                                   Data Analysis and Calculations (Con't)

the MDL (pg/L) or CRQL (pg/L) for the "A" term in Equation 4, as
appropriate.

The adjusted soil MDL or adjusted soil CRQL for all methods shall be
calculated as follows:

EQ. 6  Adjusted Soil MDL/Adjusted Soil CRQL Concentration

                                                 W    1
            Adjusted Concentration  (mg/kg) = C x — x —
                                                 WR   S
WHERE,  •     C    =   MDL or CRQL concentration (mg/kg)

             WM   =   minimum method required wet sample weight (g)

             WR   =   reported wet sample weight (g)

             S    =   % Solids/100 (see Exhibit D - Introduction to
                      Analytical Methods, Section 1.6).

For the midi-distillation, multiply the adjusted concentration value
(mg/kg) obtained in Equation 6 by any applicable dilution factor.
                            D-23/Cyanide                           ILM05.2

-------
Exhibit D  (Cyanide) — Section 12
Quality Control

12.0  QUALITY CONTROL  (QC)

12.1  Initial Calibration Verification  (ICV)

      The  ICV standard shall be prepared  in  the  same matrix as  the  calibration
      standards  and in accordance with  the instructions provided by the
      supplier.   The  ICV  standard shall be distilled.  If measurements exceed
      the  control limits  of 85%  (low) and 115%  (high), the analysis shall be
      terminated,  the  problem corrected,  the instrument recalibrated, and the
      calibration reverified.  Information regarding the ICV  shall  be reported
      on Form IIA-IN.

12.2  Continuing Calibration Verification (CCV)

      The  CCV standard shall be prepared  by  the  analyst at a  concentration
      equivalent to the mid-point of  the  calibration curve.   If the deviation
      of the CCV is greater than the  control limits of 85%  (low) and 115%
       (high), the analysis shall be stopped,  the problem corrected, the
      instrument recalibrated, the calibration verified, and  re-analysis of
      the  preceding 10 analytical samples or all analytical samples analyzed
      since the  last  compliant calibration verification shall be performed.
      Information regarding the CCV shall be reported on Form IIA-IN.

12.3  Contract Required Quantitation  Limit  (CRQL)  Check Standard  (CRI)

12.3.1   To verify linearity near the CRQL,  a standard at the CRQL  (CRI) shall
         be prepared,  in  the  same matrix  as  the  calibration standards,  and
          analyzed at  the  beginning  and  at the end of  each  sample analysis  run,
         but not before  the  ICV.   In  addition,  the Contractor shall analyze
          the CRI at a frequency  of  not  less  than once per  20  analytical
          samples1 per analysis run.   The  CRI analysis shall be run  immediately
          followed by  the  CCV  and Continuing Calibration Blank (CCB) analyses.
          The CRI shall be prepared  by spiking an aliquot of reagent water with
          cyanide to yield a  concentration in the final  solution equal to the
          CRQL.

12.3.2   CRI and percent  recovery results shall  be reported on Form IIB-IN.
          If the percent recovery falls  outside the control  limits  of 70-130%,
          the CRI shall be re-analyzed immediately.  If  the  result  of the re-
          analysis falls within the  control  limits, no further corrective
          action is required.   If the  result of the re-analysis does not fall
          within the control  limits,  the analysis shall  be  terminated,  the
          problem corrected,  the  instrument recalibrated,  the  CRI analyzed,  and
          the samples  associated with the CRI re-analyzed.

 12.4   Blank Analyses

       There are two different types of  blanks required  by  this method.   The
       calibration blank is used in  establishing the analytical curve while  the
       preparation blank is used to  monitor for possible  contamination.

 12.4.1    Initial and Continuing Calibration Blank (ICB/CCB)

          The ICB and CCB are prepared with reagents and reagent water.   If the
          absolute value of the calibration blank  (ICB/CCB)  result  exceeds the
          CRQL (see Exhibit C),  the analysis shall be terminated,  the problem
          corrected, the instrument recalibrated, the calibration verified,  and
          re-analysis of the preceding  10 analytical samples or all analytical
      •'As defined in Exhibit G,  CRI is an analytical sample.

 ILM05.2                           D-24/Cyanide

-------
                                              Exhibit D (Cyanide) — Section 12
                                                        Quality  Control  (Con't)

         samples analyzed since the last compliant calibration blank shall be
         performed.

12.4.2   Preparation Blank  (PB)

12.4.2.1    The PB shall contain all the reagents and in the same volumes as
            used in processing the samples.  The PB shall be carried through
            the complete procedure and contain the same concentration in the
            final  solution  as the sample solution used for analysis.

12.4.2.2    At least one PB, consisting of reagent water processed through
            each sample preparation and analysis procedure  (see  Section  10),
            shall  be prepared and analyzed with every Sample Delivery Group
             (SDG), or with  each batch 2 of  samples distilled, whichever  is more
            frequent.

12.4.2.3    The first batch of samples in an SDG is to be assigned to
            Preparation Blank one, the second batch of samples to Preparation
            Blank  two, etc.  (see Form III-IN).  Each Sample Data Package shall
            contain the results of all the PB analyses associated with the
            samples in that SDG.

12.4.2.4    The PB is to be reported for each SDG and used in all analyses to
            ascertain whether sample concentrations reflect contamination in
            the following manner:

12.4.2.4.1     If  the absolute value of the concentration of the blank is less
               than or equal to the CRQL  (see Exhibit C), no further action is
               required.

12.4.2.4.2     If  the analyte concentration in the blank is above the CRQL,
               the lowest concentration of the analyte in the associated
               samples shall be greater than or equal to 10 times the blank
               concentration.  Otherwise,  all samples associated with the
               blank, with  the analyte concentration less than 10 times  the
               blank concentration and above the CRQL, shall be  redistilled
               and re-analyzed with appropriate new QC.  The only exception to
               this shall be an identified field blank.  The sample
               concentration is not to be  corrected for the blank value.

12.4.2.4.3     If  the concentration of the blank is below the negative CRQL,
               then all samples associated with the blank and reported below
               10  times CRQL shall be reprepared and re-analyzed with
               appropriate  new QC.

               The values for  the preparation blank shall be reported on Form
               III-IN.

12.5  Spike Sample Analysis

12.5.1   The spike sample analysis is designed to provide information about
         the effect of the  sample matrix on the distillation and/or
         measurement methodology.  The spike is added prior to any
         distillation steps.  At least one spike sample analysis  (matrix
         spike) shall be performed on each group of samples of a similar
      2A group of samples prepared at the same time.

                                ,  D-25/Cyanide                           ILM05.2

-------
Exhibit D  (Cyanide) — Section 12
Quality Control  (Con't)

         matrix  type  (i.e., water,  soil)  or  for each SDG. 3  The sample and its
         associated spike  sample  shall  initially be run at the  same dilution.

12.5.2   If the  spike  analysis is performed  on the same sample  that is chosen
         for  the duplicate sample analysis,  spike calculations  shall be
         performed using the  results  of the  sample designated as the "original
         sample" (see  Section 12.6).  The average of the duplicate results
         cannot  be used for the purpose of determining percent  recovery.
         Samples identified as field  blanks  and Performance Evaluation  (PE)
         samples shall not be used  for  spiked sample analysis.  USEPA may
         require that  a specific  sample be used for the spike sample analysis.

12.5.3   The  analyte spiking  solution shall  be added to yield a final
         concentration of  100 pg/L  in the final sample solution prepared for
         analysis (i.e., post-distillation).  The final volume  of the sample
         after distillation shall be  the basis for the amount of cyanide to be
         added as the  spike.   For instance,  the full volume distillation
         procedure will require addition of  25 ug cyanide to  the sample prior
         to distillation  [based on  the  final distillate volume  of 250
         milliliter  (mL)]  to  meet the specified spiking level;  and the midi
         distillation  procedure requires the addition of 5 ug of cyanide to
         the  sample prior  to  distillation (based on the final distillate
         volume  of 50  mL).

12.5.3.1    For  soil samples,  the final sample solution prepared for analysis
             (i.e., the distillate)  shall contain cyanide spiked at a
             concentration  of  100  ug/L regardless of the distillation procedure
             employed,  or  the  amount of  sample used  for distillation.  The
             final sample  volume after distillation  shall be used as  the basis
             for  the  amount of cyanide to add as the spike.  The units for
             reporting  soil sample cyanide results  shall be mg/kg.  To convert
             from pg/L  to  mg/kg, the equation below  shall be used:
                                                          -1
             EQ.  7 Conversion to  mg/kg


                                  final distillate volume (L)
                    mg/kg = ug/L x
                                       sample weight (g)
 12.5.4    If the spike recovery is not at or within the limits  of 75-125%,  the
          data of all samples received and associated with that spike sample
          and determined by the same analytical method shall be flagged with
          the letter "N" on Forms IA-IN and VA-IN.   An exception to this rule
          is granted when the sample concentration exceeds the  spike added
          concentration by a factor of four or more.  In such an event,  the
          data shall be reported unflagged even if the percent  recovery does
          not meet the 75-125% recovery criteria.

 12.5.5    When the matrix spike recovery falls outside the control limits and
          the sample result does not exceed 4 times the spike added, a post-
          distillation spike shall be performed.  Note that if  a post-
          distillation spike analysis is required,  the same USEPA sample that
          was used for the matrix spike analysis shall be used  for the post
          digestion spike analysis.  Spike the unspiked aliquot of the sample
          at 2 times the indigenous level or 2 times CRQL, whichever is
          greater.  Results of the post-distillation spike shall be reported on
          Form VB-IN.
       3USEPA may require additional  spike  sample  analyses, upon USEPA Regional
 CLP Project Officer (CLP PO)  request.

 ILM05.2                          D-26/Cyanide

-------
                                              Exhibit D (Cyanide) — Section 12
                                                        Quality Control  (Con't)

12.5.6   In the instance where there is more than one spike sample per matrix,
         per method, per SDG, if one spike sample recovery is not within
         contract criteria, flag all the samples of the same matrix and method
         in the SDG.  Individual component percent recoveries are calculated
         as follows:

         EQ. 8  Spike Percent Recovery

                           % Recovery =  SSR " SR x 100
                                           SA
         WHERE,
                            1
          SSR   =   Spiked Sample Result

          SR    =   Sample Result

          SA    =   Spike Added

12.5.7   When the sample concentration is less than the Method Detection Limit
          (MDL), use SR  = 0 only for purposes of calculating percent recovery.
         The Spike Sample Results  (SSRs), Sample Results  (SRs), Spike Added
          (SA),  and percent recovery  (positive or negative) shall be reported
         on Form VA-IN.

12.5.8   The units used for reporting spike sample results will be identical
         to those used  for reporting sample results on Form IA-IN.

12.6  Duplicate Sample Analysis

12.6.1   One duplicate  sample shall be analyzed from each group of samples of
         a similar matrix type  (i.e., water, soil) or for each SDG. 4
         Duplicates cannot be averaged for reporting on Form IA-IN.  The
         sample and its associated duplicate sample shall initially be run at
         the same dilution.

12.6.2   Duplicate sample analyses are required for percent solids.  Samples
         identified as  field blanks and  PE samples shall not be used for
         duplicate sample analysis.  USEPA may require that a specific sample
         be used for duplicate  sample analysis.  The Relative Percent
         Difference  (RPD) is calculated  as follows:

         EQ. 9  Duplicate Sample Relative Percent Difference


                              RPD =  I S " D I x 100
                                     (S+D)/2


         WHERE,

          RPD   =   Relative  Percent Difference

          S     =   Sample Result (original)

          D     =   Duplicate  Result

12.6.3   The results of the duplicate sample analyses shall be reported on
         Form VI-IN.  A control  limit of 20% for RPD shall be used for
      4USEPA may require additional duplicate sample analyses,  upon USEPA
Regional CLP PO request.

                                  D-27/Cyanide                           ILM05.2

-------
Exhibit D  (Cyanide) — Section 12
Quality Control  (Con't)

         original  and  duplicate  sample values greater  than or equal  to  five
         times the CRQL  (see Exhibit C).  A control limit of the  CRQL value
         shall be  entered  in the "Control Limit" column on Form VI-IN if
         either  the sample or duplicate value is less  than five times the
         CRQL.   If the sample and duplicate values are greater than  or  equal
         to  five times the CRQL,  or if the sample and  duplicate values  are
         less than the CRQL, the "Control Limit" field is left empty.

12.6.4   If  one  result is  above  five times the CRQL level and the other is
         below,  use the  CRQL criteria to determine if  the duplicate  analysis
         is  in control.  If both sample and duplicate  values are  less than the
         MDL, the  RPD  is not calculated on Form VI-IN.  For  solid sample or
         solid duplicate results less than five times  the CRQL, enter the
         value of  the  CRQL, corrected for sample weight and  percent  solids,
          (i.e.,  original,  not duplicate sample weight  and percent solids), in
         the "Control  Limit" column.  If the duplicate sample results are
         outside the control limits, flag all the data for samples received
         and associated  with that duplicate sample with an "*" on Forms IA-IN
         and VI-IN.  In  the instance where there is more than one duplicate
         sample  per SDG, if one  duplicate result is not within contract
         criteria, flag  all samples of  the same matrix and method in the SDG.
         The percent difference' data will be used by USEPA to evaluate  the
         long-term precision of  the method.  Specific  control limits for each
         element will  be added  to Form VI-IN at a later date based on the
         precision results.

12.7   Laboratory Control Sample  (LCS) Analysis

12.7.1  A solid LCS  (LCSS) shall be analyzed using the same sample
         preparations, analytical methods, and Quality Assurance  (QA)/QC
         procedures employed  for the EPA  samples received.   For  cyanide, a
         distilled ICV shall be  used as the aqueous LCS  (LCSW).

12.7.2  The USEPA provided LCSS shall be prepared and analyzed using each of
         the procedures  applied to the  solid  samples  received  (exception:
         percent solids  determination  not  required).   If  the USEPA LCSS is
         unavailable,  other USEPA QC Check  samples  or  other  certified
         materials may be  used.   In such  a  case,  the  control limits  for LCSS
         must be documented and provided.   One  LCSS  shall be prepared and
         analyzed for  every group of  solid  samples  in  a  SDG,  or  for  each batch
         of samples distilled,  whichever  is more  frequent.

12.7.3  All LCSS  and  percent  recovery  results will be reported  on Form VII-
          IN.  If the  results  for the  LCSS  fall  outside the  control  limits
          established  by USEPA,  the analyses  shall be  terminated,  the problem
          corrected,  and the samples associated with that  LCSS  reprepared and
          re-analyzed  with  appropriate  new QC.

12.8   Method Detection Limit  (MDL) Determination

12.8.1  Before  any  field  samples are  analyzed  under  this  contract,  the MDLs
          shall be determined for non-distilled analyses  (Preparation
          Method/Code  "NP1") and for each  distillation procedure  and instrument
          used,  prior  to the start of the  contract  analyses,  and annually
          thereafter,  and  shall meet the levels  specified in Exhibit  C.

          An MDL study shall be performed after major instrument  maintenance,
          or changes  in  instrumentation or instrumental conditions to verify
          the current   sensitivity of the analysis.

 12.8.2   To determine  the  MDLs,  the Contractor  shall  run  MDL studies following
          the procedures given in 40 CFR,  Part 136.   The  Contractor shall


 ILM05.2                           D-28/Cyanide

-------
                                          Exhibit D (Cyanide)  — Sections 12-17
                                                             Method Performance

         prepare the MDL samples by each distillation procedure used and shall
         analyze these samples on each instrument used.  The Contractor shall
         also analyze the non-distilled MDL samples on each instrument used.

12.8.3   The determined concentration of the MDL shall be less than half the
         concentration of the CRQL listed in Exhibit C.

12.8.4   The non-distilled MDL (Preparation Method/Code "NP1") shall be used
         to determine the appropriate concentration qualifier for the results
         of instrument QC analyses (except the distilled ICV).

12.8.5   The results of the MDL determination study shall be forwarded to the
         USEPA Regional CLP PO, Sample Management Office (SMO), and Quality
         Assurance Technical Support  (QATS).

12.8.6   The MDL results shall be reported on Form IX-IN.

13.0  METHOD PERFORMANCE

      Not applicable.

14.0  POLLUTION PREVENTION

      See Section 1.15 in Exhibit D - Introduction to Analytical Methods.

15.0  WASTE MANAGEMENT

      See Section 1.16 in Exhibit D - Introduction to Analytical Methods.

16.0  REFERENCES

16.1  US Environmental Protection Agency. Methods for Chemical Analysis of
      Water and Wastes. Method 335.2. 1980.

16.2  American Water Works Association/American Public Health
      •Association/Water Environment Federation. Standard Methods for the
      '^Examination of Water and Wastewater. Method 4500. 18th   Edition.

16.3  US Government Printing Office.  40 Code of Federal Regulations, Part 136,
      Section 1, Appendix B.

17.0  TABLES/DIAGRAMS/FLOWCHARTS

      Not applicable.
                                  D-29/Cyanide                          ILM05.2

-------
                          EXHIBIT E




CONTRACT LABORATORY PROGRAM QUALITY ASSURANCE MONITORING PLAN
                              E-l                                ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               E-2

-------
  Exhibit E - Contract Laboratory  Program Quality Assurance Monitoring Plan

                               Table of Contents

Section                                                                    Page


1.0   OVERVIEW   	5
      1.1   Quality Assurance/Quality Control  (QA/QC) Activities    	   5
      1.2   Incentives/Sanctions   	   5

2.0   INTRODUCTION   	   6
      2.1   Quality Assurance/Quality Control  (QA/QC) Program Components  .   6

3.0   GENERAL QUALITY ASSURANCE/QUALITY CONTROL (QA/QC)  REQUIREMENTS  ...   7

4.0   SPECIFIC QUALITY ASSURANCE/QUALITY CONTROL (QA/QC)
      MONITORING PROCEDURES 	   8
      4.1   Purpose	8
      4.2   Laboratory Audit and Intercomparison Study Program    	   8
      4.3   Annual Verification of Method Detection Limits (MDLs) 	   8
      4.4   Quarterly Verification of Linear Ranges/Interelement
            Correction Factors   	   9
      4.5   Quality Assurance/Quality Control Measurements   	   9

5.0   QUALITY ASSURANCE MANAGEMENT PLAN 	 10
      5.1   Introduction   	10
      5.2   Required Elements of a Quality Assurance Management Plan    ...  10
      5.3   Updating and Submitting the Quality Assurance
            Management Plan	12
      5.4   Incentives/Sanctions   	  13

6.0   STANDARD OPERATING PERFORMANCE STANDARDS   	  14
      6.1   Introduction   	14
      6.2   Format   	15
      6.3   Required SOPs	15
      6.4   Updating and Submitting SOP Requirements   	18
      6.5   Incentives/Sanctions   	  19

7.0   CONTRACT COMPLIANCE SCREENING (CCS)  PERFORMANCE STANDARDS 	  20
      7.1   Overview   	20
      7.2   CCS Results	20
      7.3   CCS Trend Report   	20
      7.4   Incentives/Sanctions   	  20

8.0   ANALYTICAL PERFORMANCE STANDARDS REQUIREMENT S 	  21
      8.1   Overview   	21
      8.2   Preparation of Chemical Standards from the Neat
            High Purity Bulk Materia 1	21
      8.3   Purchase of Chemical Standards Already in Solution    	 21
      8.4   Requesting Standards from the USEPA Standards
            Repository	24
      8.5   Documentation of the Verification and
            Preparation of Chemical Standards  	 24
      8.6   Incentives/Sanctions   	  25

9.0   DATA PACKAGE MONITORING AUDITS   	  26
      9.1   Overview   	26
      9.2   Responding to the Data Package Audit Repor t	26
      9.3   Incentives/Sanctions   	  26

10.0  REGIONAL DATA REVIEW MONITORING 	  27
      10.1  Overview   	27
                                      E-3                               ILM05.2

-------
  Exhibit  E  - Contract  Laboratory Program Quality Assurance Monitoring  Plan

                           Table of Contents (Con't)

Section                                                                    Page


11.0  QUALITY ASSURANCE  (QA) PROFICIENCY MONITORIN G 	 28
      11.1  Performance Evaluation  (PE) Samples  	  28
      11.2  Quarterly Blind  (QB) Audits 	  28
      11.3  Incentives/Sanctions   	  30

12.0  ON-SITE LABORATORY QUALITY ASSURANCE  (QA) MONITORING EVALUATIONS  . . 31
      12.1  Overview   	31
      12.2  Quality Assurance On-Site Evaluation    	 31
      12.3  Evidentiary Audit  	  31
      12.4  Discussion of the On-Site Team's Finding s  	 32
      12.5  Incentives/Sanctions   	 32

13.0  ELECTRONIC DATA QUALITY ASSURANCE (QA) MONITORING AUDITS    	 33
      13.1  Overview   	33
      13.2  Submission of the Instrument  Electronic Data   	35
      13.3  Responding to the Electronic  Data Audit Report    	 35
      13.4  Incentives/Sanctions   	 35

14.0  DATA MANAGEMENT PERFORMANCE REQUIREMENTS   	 36
      14.1  Overview   	36
      14.2  Documenting  Data Changes   	36
      14.3  Lifecycle Management  Procedures  	  36
      14.4   Personnel Responsibilities    	 37

15.0  TABLES   	38
      TABLE  1. Contract  Laboratory  Program Quality Assurance
               Monitoring Plan    	38
 ILM05.2                                E-4

-------
                                                         Exhibit E — Section 1
                                                                        Overview
1.0   OVERVIEW
      Quality Assurance  (QA) and Quality Control  (QC) are integral parts of
      the U.S. Environmental Protection Agency's  (USEPA's) Contract Laboratory
      Program (CLP).   The QA process consists of management review and
      oversight at the planning, implementation, and completion stages of the
      environmental data collection activity, and ensures that data provided
      are of the quality required.  The QC process includes those activities
      required during data collection to produce the data quality desired and
      to document the quality of the collected data.

1.1   Quality Assurance/Quality Control (QA/QC) Activities

      During the planning of an environmental data collection program, QA
      activities focus on defining data quality criteria and designing a QC
      system to measure  the quality of data being generated.  During the
      implementation of  the data collection effort, QA activities ensure that
      the QC system is functioning effectively, and that the deficiencies
      uncovered by the QC system are corrected.  After environmental data are
      collected, QA activities focus on assessing the quality of data obtained
      to determine its suitability to support enforcement or remedial
      decisions.

1.1.1    This exhibit describes the overall QA/QC operations and the processes
         by which the CLP meets the QA/QC objectives defined above.  This
         contract requires a variety of QA/QC activities.  These contract
         requirements are the minimum QC operations necessary to satisfy the
         analytical requirements associated with the determination of the
         different method analytes.  These QC operations are designed to
         facilitate laboratory comparison by providing USEPA with comparable
         data from all Contractors.  These requirements do not release the
         analytical Contractor from maintaining their own QC checks on method
         and instrument  performance.

1.2   Incentives/Sanctions

      The Contractor may anticipate incentives by consistently providing the
      following:  (1)  high quality, technically sound data as stipulated by the
      ILM05.2 contract;  (2) on-time or early delivery of the Sample Delivery
      Group  (SDG) Cover  Sheet;  (3) above average Quarterly Blind (QB)
      Performance Evaluation  (PE) sample scores;  (4) diskettes that pass the
      initial Contract Compliance Screening  (CCS) acceptance criteria; and (5)
      SDGs delivered on-time.  Samples are distributed routinely to
      Contractors based  on the quality of work performed, as measured by the
      Performance Scheduling Algorithm  (PSA)  (see Section G of the contract
      for details).  A Contractor that consistently meets the contract
      performance requirements as highlighted above, will earn a higher PSA
      score, thereby increasing the likelihood of receiving samples for
      analyses.  If the  Contractor fails to meet the requirements set forth in
      this Statement of  Work  (SOW) or elsewhere in the contract, USEPA may
      take, but is not limited to, the following actions  (see Section E of the
      contract  for details): reduction in the number of samples sent under the
      contract; suspension of sample shipments; data package audit(s);
      electronic data audit(s); on-site laboratory evaluation(s); and/or
      remedial  PE sample(s).
                                       E-5                                ILM05.2

-------
Exhibit E — Section 2
Introduction

2.0   INTRODUCTION

      Appropriate use of data generated under the large range of analytical
      conditions encountered in environmental analyses requires reliance on
      the Quality Control  (QC) procedures and criteria incorporated  into the
      ILM05.2 Statement of Work  (SOW).

      The data acquired from QC procedures are used  to estimate and  evaluate
      the information content of analytical  results  and to determine the
      necessity for, or the effect of, corrective action procedures.  The
      parameters used to estimate information content include precision,
      accuracy, detection  limit, and other quantitative and qualitative
      indicators.   In addition, QC procedures give an overview of  the
      activities required  in an integrated program to generate data  of known
      and documented quality required  to meet defined objectives.

2.1   Quality Assurance/Quality Control  (QA/QC)  Program Components

2.1.1    The Contractor's  QA/QC program shall include  (1) internal QC criteria
         that demonstrate  compliant levels of performance, as  determined by QA
         review, as well as  (2)  external  review  of data and procedures
         accomplished by the monitoring activities of the USEPA OERR
         Analytical Operations/Data Quality  Center  (AOC), Regional Data Users,
         Sample Management Office  (SMO),  and the Quality Assurance Technical
         Support  (QATS) Laboratory.  Each external review accomplishes a
         different  purpose.  These  reviews are described in specific sections
         of this exhibit.  Laboratory  evaluation samples, electronic data
         audits, and  data  packages  provide an external QA reference  for the
         program.   A  Contractor  on-site evaluation system is  also part of  the
         external QA  monitoring.  A feedback loop provides the results of  the
         various review functions to the  Contractors through  direct
         communications with  the USEPA Regional  Contract Laboratory  Program
         Project Officer  (CLP  PO) and  the USEPA  OERR AOC Inorganic Program
         Manager  (AOC PM).

 2.1.2    This exhibit does not provide specific  instructions  for  constructing
         QA Management  Plans,  QC systems, or a QA organization.   It  is,
         however, an  explanation of the QA/QC  requirements of CLP.  It
         outlines minimum standards for  QA/QC programs.  It also  includes
         specific  items that  are required in a Quality Assurance  Management
         Plan  (QAP) and by the QA/QC documentation detailed  in this  contract.
         Delivery of  this  documentation  provides USEPA with a complete data
         package which will  stand alone,  and limits  the  need  for  contact  with
         the Contractor or with  an  analyst,  at  a later date,  if some aspect  of
         the analysis is  questioned.

 2.1.3    In order to  assure  that the product delivered by the Contractor meets
         the requirements  of the contract,  and  to improve  interlaboratory data
         comparison,  the  Contractor shall:

         ="   Prepare,  and adhere to,  a written  approved QAP,  as  defined in
              Exhibit E, Section 5;

          :x-   Prepare and  adhere to, Standard Operating Procedures (SOPs)  as
              described in Exhibit  E,  Section 6;

          '•*•   Adhere  to the analytical methods in Exhibit D and associated QC
              requirements specified within  Exhibit  E;

          '•*'•   Verify  and document analytical standards and retain
              documentation of the  purity of neat materials,  as well as, the
 ILM05.2                               E-6

-------
                                                    Exhibit E — Sections 2 & 3
                                                     General QA/QC Requirements

              purity and accuracy of solutions obtained from private chemical
              supply houses;

         •*•    Submit all raw  data and required documentation for Regional
              review;

         ••*=    Submit results  of all analyzed laboratory evaluation samples,
              and adhere to corrective action procedures;

         •*••    Submit,  upon request, instrument data tapes and applicable
              documentation for tape audits, including a copy of the Sample
              Data Package;

         '•*'    Submit to on-site laboratory evaluations, and adhere to
              corrective action procedures; and

         ••ť    Submit all original documentation generated during sample
              analyses for USEPA review.

3.0   GENERAL QUALITY ASSURANCE/QUALITY CONTROL  (QA/QC) REQUIREMENTS

      The Contractor shall adhere  to USEPA's Good Laboratory Practices for
      laboratory cleanliness with  regard to glassware  and apparatus.  The
      Contractor shall also adhere to good  laboratory  practices with regard to
      reagents, solvents, and gases.  For additional guidelines regarding
      these general laboratory procedures,  see the Handbook for Analytical
      Quality Control in Water and Wastewater Laboratories USEPA-600/4-79-019.
      USEPA Environmental Monitoring Systems Laboratory, Cincinnati, Ohio,
      September 1982.
                                      E-7                                ILM05.2

-------
Exhibit E — Section 4
Specific QA/QC Monitoring Procedures

4.0   SPECIFIC QUALITY ASSURANCE/QUALITY CONTROL  (QA/QC) MONITORING PROCEDURES

4.1   Purpose

4.1.1    The purpose of this document is to provide  (1) a uniform set of
         procedures for the analysis of inorganic constituents of samples,  (2)
         documentation of methods  and their performance, and  (3) verification
         of the sample data generated.  Although  it  is  impossible to address
         every analytical situation in one document, this exhibit defines the
         minimum requirements  for  each major step relevant to any inorganic
         analysis.

4.1.2    The primary function  of the Contract Laboratory Program  (CLP) QA/QC
         program is the definition of procedures  for the evaluation and
         documentation of analytical methodologies and  the reduction and
         reporting of data.  The location and summary of the QA/QC performance
         based contracting methods can be found in Exhibit E, Section 15,
         Table 1 - Contract Laboratory Program Quality  Assurance Monitoring
         Plan.  The objective  is to provide a uniform basis for sample
         handling, instrument  and  methods maintenance,  performance evaluation,
         and analytical data gathering and reporting.   In many instances where
         methodologies are available, specific QC procedures are incorporated
         into the method documentation  (see Exhibit  D).

4.1.3    The QA/QC procedures  defined herein shall be used by the Contractor
         when performing the methods specified in Exhibit D.  When QA/QC
         procedures are specified  in Exhibit D, the  Contractor shall  follow
         those procedures, in  addition to procedures specified here.

4.2    Laboratory Audit and Intercomparison Study  Program

       The  Contractor  is required to participate in the  Laboratory Audit and
       Intercomparison  Study  Program run by USEPA.  The  Contractor shall be
       required to analyze  at  least one Quarterly  Blind  (QB) sample per
       calendar quarter during  the  contract period for  inorganics.

4.3    Annual Verification  of Method Detection Limits (MDLs)

       The  Contractor  shall perform and report annual verification of  MDLs by
       the  method  specified in  Exhibit D, by type, matrix, and model for each
       instrument  used on  this  contract, to Sample Management Office  (SMO),
       Quality Assurance Technical  Support  (QATS), and  the USEPA Regional
       Contract Laboratory  Program  Project Officer (CLP  PO) as specified in
       Exhibit B.  All  the  MDLs shall meet the requirements specified  in
       Exhibit C.
 ILM05.2                               E-8

-------
                                                         Exhibit E — Section 4
                                   Specific QA/QC Monitoring Procedures  (Con't)


4.4   Quarterly Verification of Linear Ranges/Interelement Correction Factors

      The Contractor shall perform and report quarterly verification of linear
      ranges by the method specified in Exhibit D, by type and model for each
      instrument used on this contract, to SMO, QATS, and the USEPA Regional
      CLP PO as specified in Exhibit B.  The Contractor shall also report, as
      specified in Exhibit B, integration times.  For Inductively Coupled
      Plasma - Atomic Emission Spectroscopy  (ICP-AES) methods, the Contractor
      shall also report, as specified in Exhibit B, wavelengths used and all
      interelement correction factors.

4.5   Quality Assurance/Quality Control Measurements

4.5.1    In this Exhibit, as well as other places within this Statement of
         Work  (SOW), the term "analytical sample" discusses the required
         frequency or placement of certain QA/QC measurements.  The term
         "analytical sample" is defined in the glossary, Exhibit G.

4.5.2    In order for the QA/QC information to reflect the status of the
         samples analyzed, all samples and their associated QA/QC analysis
         shall be analyzed under the same operating and procedural conditions.

4.5.3    If any QC measurement fails to meet contract criteria, the analytical
         measurement must not be repeated prior to•taking the appropriate
         corrective action as specified in Exhibit D.  The exception is the
         CRI analysis, which may be re-analyzed once before corrective action
         is necessary.

4.5.4    The Contractor shall report all QC data in the exact format specified
         in Exhibits B and H.

4.5.5    MDLs, precision, linear dynamic range, and interference effects shall
         be established for each analyte on a particular instrument.  All
         reported measurements shall be within the instrumental linear ranges.
         The Contractor shall maintain QC data confirming instrument
         performance and analytical results.
                                      E-9                                ILM05.2

-------
Exhibit E — Section 5
QA Management Plan

5.0   QUALITY ASSURANCE MANAGEMENT PLAN  (QAP)

5.1   Introduction

      The Contractor shall establish a Quality Assurance  (QA) program with the
      objective of providing  sound .analytical chemical measurements.  This
      program shall incorporate the Quality  Control  (QC)  procedures, any
      necessary corrective action, all documentation required during data
      collection, and the quality assessment measures performed by management
      to ensure acceptable data production.  The Contractor  shall follow the
      USEPA EPA Requirements  for Quality Management Plans  (EPA QA/R-2).  An
      electronic version can  be found at http://www.epa.gov/qualityl/
      qa_docs.html.

5.1.1    The Contractor shall prepare a  written QAP which describes the
         procedures that are  implemented to  achieve  the  following:

         30  Maintain data  integrity, validity,  and  usability;

         :x:  Ensure that analytical  measurement  systems  are  maintained in  an
             acceptable  state of  stability and  reproducibility;

         00  Detect problems  through data  assessment and establish  corrective
             action procedures which keep  the analytical process  reliable;  and

         '•*•  Document all  aspects  of the measurement process in  order  to
             provide data  which are  technically sound and legally defensible.

5.1.2    The QAP must present,  in  specific terms,  the  policies, organization,
         objectives, functional guidelines,  and specific QA/QC activities
         designed  to achieve  the  data quality requirements in this  contract.
         Standard  Operating Procedures  (SOPs) pertaining to  each  element shall
         be included or  referenced as part of the  QAP.   The  QAP  shall  be
         paginated consecutively  in  ascending order.   The QAP shall be
         available during  on-site  laboratory evaluations and shall  be
         submitted to the  designee within  7  days of  written  request by the
         USEPA Regional  Contract  Laboratory  Program  Project  Officer (CLP PO)
         or the USEPA OERR Analytical Operations/Data Quality Center (AOC)
         Inorganic Program Manager (AOC PM). Additional information relevant
         to the preparation of a  QAP can be  found in USEPA and ASTM
         publications.

5.2    Required Elements  of  a  Quality Assurance  Management Plan

       The  required elements of a  laboratory's QAP  are outlined in this
       section.   This outline  shall be used as a framework for developing the
       QAP.

       A.   Organization  and Personnel

          1.  QA Policy  and Objectives (the mission and quality policy of the
              organization)

          2.  QA Management (the specific roles,  authorities, and
              responsibilities of management  and staff with respect to QA and
              QC activities)

             a.  Organization

             b.  Assignment of QA/QC Responsibilities
 ILM05.2                               E-10

-------
                                                   Exhibit E — Section 5
                                               QA Management Plan  (Con't)

      c.  Reporting Relationships  (the means by which effective
          communications with personnel actually performing the work are
          assured)

      d.  QA Document Control Procedures

      e.  QA Program Assessment Procedures  (the process used to plan,
          implement, and assess the work performed)

   3.  Personnel

      a.  Resumes

      b.  Education and Experience Pertinent to this Contract

      c.  Training Records and Progress

B.  Facilities and Equipment

   1.  Instrumentation and Backup Alternatives

   2.  Maintenance Activities and Schedules

C.  Document Control

   1.  Laboratory Notebook Policy

   2.  Sample Tracking/Custody Procedures

   3.  Logbook Maintenance and Archiving Procedures

   4.  Sample Delivery Group  (SDG) File Organization, Preparation, and
       Review Procedures

   5.  Procedures for Preparation, Approval, Review, Revision, and
       Distribution of SOPs

   6.  Process for Revision of Technical or Documentation Procedures

D.  Analytical Methodology

   1.  Calibration Procedures and Frequency

   2.  Sample Preparation Procedures

   3.  Sample Analysis Procedures

   4 .  Standards Preparation Procedures

   5.  Decision Processes, Procedures, and Responsibility for Initiation
       of Corrective Action

E.  Data Generation

   1.  Data Collection Procedures

   2.  Data Reduction Procedures

   3.  Data Validation Procedures

   4.  Data Reporting and Authorization Procedures



                                E-ll                               ILM05.2

-------
Exhibit E — Section 5
QA Management Plan  (Con't)

      F.  Quality Assurance  (the process which measures the effectiveness of
          QA will be established and how frequently effectiveness  will be
          measured)

         1.  Data Quality Assurance

         2.  Systems/Internal  Audits

         3.  Performance/External  Audits

         4.  Corrective  Action Procedures  (the continual  improvement based  on
             lessons learned from  previous  experience)

         5.  QA Reporting Procedures

         6.  Responsibility  Designation

      G.  Quality Control

         1.  Solvent,  Reagent, and Adsorbent Check Analysis

         2.  Reference Material Analysis

         3.  Internal  QC Checks

         4.  Corrective  Action and Determination of QC Limit  Procedures

         5.  Responsibility  Designation

 5.3   Updating  and  Submitting the  Quality Assurance Management Plan

 5.3.1    The revised QAP will become  the official QAP under the contract and
         may be used during  legal  proceedings.   The Contractor shall maintain
         the QAP on file at  the Contractor's facility for the term of the
         contract.  Both the initial  submission  and the revised QAP shall be
         paginated consecutively  in  ascending order.   The revised QAP shall
         include:

         ^   Changes  resulting from (1) the Contractor's internal review of
              their organization,  personnel, facility, equipment, policy and
              procedures, and  (2)  the Contractor's implementation of .the
              requirements of the contract,  and

         =*•   Changes  resulting from USEPA' s review of the laboratory
              evaluation sample data,  bidder supplied documentation, and
              recommendations made during the pre-award on-site laboratory
              evaluation.

 5.3.1.1     The Contractor shall  send a copy of  the  latest version of the QAP
             within 7  days of a request from a USEPA Regional  CLP PO or the
             USEPA OERR AOC PM.  The request will designate the recipients.

 5.3.2    Subsequent Updates  and Submissions.  During  the term of the contract,
          the Contractor shall amend the QAP when the  following circumstances
          occur:

          •*•   USEPA modifies the technical requirements of the Statement of
              Work  (SOW)  or contract;

          ••*•   USEPA notifies the Contractor' of deficiencies in the QAP
              document;
 ILM05.2                               E-12

-------
                                                         Exhibit E — Section 5
                                                     QA Management Plan  (Con't)

         ^   USEPA notifies the Contractor of deficiencies resulting from
             USEPA's review of the Contractor's performance;

         :x:   The Contractor's organization, personnel, facility, equipment,
             policy, or procedures change; or

         •*=   The Contractor identifies deficiencies resulting  from the
             internal review of their organization, personnel,  facility,
             equipment, policy, or procedures changes.

5.3.2.1     The Contractor shall amend the QAP within 14 days of when the
            circumstances listed in Exhibit E, Section 5.3, result in a
            discrepancy between what was previously described  in the QAP and
            what is presently occurring at the Contractor's facility.  When
            the QAP is amended, all changes in the QAP shall be clearly marked
             (e.g., a bar in the margin indicating where the change is found in
            the document, highlighting the change by underlining the change,
            bold printing the change, or using a different print font) and a
            copy is sent to the USEPA Regional CLP PO and Quality Assurance
            Technical Support  (QATS).  The amended section pages shall have-
            the date on which the changes were implemented.  The Contractor
            shall incorporate all amendments to the latest version of the QAP
            document.  The Contractor shall archive all amendments to the QAP
            document for future reference by USEPA.

5.4   Incentives/Sanctions

      The Contractor shall amend the QAP as specified within this section.
      The QAP describes the policies and procedures for ensuring that work
      processes, products, or services satisfy expectations or  specifications
      in ILM05.2.  Failure to comply with the requirements of  this section may
      result in sanctions as described in the contract.
                                      E-13                               ILM05.2

-------
Exhibit E — Section 6
Standard Operating Performance Standards

6.0   STANDARD OPERATING PERFORMANCE STANDARDS

6.1   Introduction

      In order to obtain reliable  results,  adherence  to  prescribed analytical
      methodology is imperative.   In any  operation  that  is  performed on  a
      repetitive basis, reproducibility is  best  accomplished through the use
      of Standard Operating Procedures  (SOPs).   As  defined  by USEPA,  an  SOP is
      a written document which provides directions  for the  step-by-step
      execution of an  operation, analysis,  or  action  which  is commonly
      accepted as the  method for performing certain routine or repetitive
    N tasks.  The Contractor shall follow the  USEPA Guidance for the
      Preparation of Standard Operating Procedures  (SOPs) for Quality-Related
      Documents  (EPA QA/G-6).  An  electronic version  can be found at
      http://www.epa.gov/qualityl/qa_docs.html.

6.1.1    SOPs prepared by  the Contractor  shall be functional (i.e.,  clear,
         comprehensive, up-to-date,  and sufficiently  detailed to permit
         duplication of results by qualified analysts).   The SOPs shall  be
         paginated  consecutively in ascending order.

6.1.2    All  SOPs shall reflect Contractor  activities as they are currently
         performed  in  the  laboratory.   In addition, all  SOPs shall be:

         '•**   Consistent with current USEPA regulations,  guidelines, and the
              Contract  Laboratory  Program (CLP) ILM05.2 contract requirements.

         '•^   Consistent with instrument (s)  manufacturer's specific instruction
              manuals.

         :*'   Available to USEPA during an on-site laboratory evaluation.  A
              complete set of SOPs  shall be bound together and available for
              inspection at such evaluations.  During on-site laboratory
              evaluations,  laboratory personnel shall demonstrate the
              application of the SOPs if requested.

         :x:   Available to the designated recipients within 7 days, upon
              request by the USEPA Regional CLP Project Officer  (CLP PO)  or the
              USEPA OERR Analytical Operations/Data Quality Center  (AOC)
              Inorganic Program Manager (AOC PM).

          ^   Capable of providing for the development of documentation that is
              sufficiently complete to record the performance of all tasks
              required by the protocol.

          =*•   Capable of demonstrating the validity of data reported by the
              Contractor and explain the cause of missing or inconsistent
              results.

          :x:   Capable of describing the corrective measures and  feedback
              mechanism utilized when analytical results do not meet protocol
              requirements.

          :*:   Reviewed regularly and updated as necessary when contract,
              facility, or Contractor procedural modifications are made.

          =*=   Archived for  future  reference in usability or evidentiary
              situations.

          :*"   Available at  specific work stations as appropriate.
 ILM05.2                               E-14

-------
                                                          Exhibit E — Section 6
                               Standard Operating Performance Standards  (Con't)

         •*••   Subject to  a document  control  procedure which precludes the use
             of outdated or  inappropriate SOPs.

         •*•   Reviewed and signed by all  Contractor  personnel performing
             actions identified in  the SOP.

6.2   Format

      The format for SOPs may vary  depending  upon the type of activity for
      which they are prepared; however,  at  a  minimum,  the following sections
      shall be included:

      ="   Title page;

      :xl   Document Control;

      '•*>   Scope and Applicability;

      '•**   Summary of Method;

      :ť   Definitions  (acronyms, abbreviations,  and specialized forms used in
          the SOP);

      ="   Health & Safety;

      ;*'   Personnel Qualifications;

      *"   Interferences;

      =*>   Apparatus & Materials  (list or specify; note also designated
          locations where found);

      '"   Handling & Preservation;

      ^   Instrument or  Method Calibration;

      -!3tl   Sample Preparation and Analysis;

      00   Data Calculations;

      '**>   Quality Control  (QC) limits;

      111   Corrective action  procedures,  including procedures for secondary
          review of information being generated;

      =*>   Data Management and Records Management;

      :xl   Miscellaneous  notes and precautions;  and

      111   References.

6.3   Required SOPs

      The Contractor shall maintain the  following SOPs:

6.3.1    Evidentiary SOPs for required chain-of-custody and document control
         are discussed in Exhibit F.
                                      E-15                               ILM05.2

-------
Exhibit E — Section  6
Standard Operating Performance  Standards  (Con't)

6.3.2    Sample Receipt  and Storage

         !M!  Sample  receipt and identification logbooks,

         '•*>  Refrigerator temperature logbooks, and

         ^  Security precautions.

6.3.3 Sample Preparation

6.3.3.1     Metals

6.3.3.2     Cyanide

6.3.4    Glassware Cleaning

6.3.5    Calibration (Balances, etc.)

         -JC   Procedures;

         :*:   Frequency requirements;

         ^   Preventative maintenance schedule and procedures;

         00   Acceptance criteria and corrective actions; and

         '"'   Logbook maintenance authorization.

6.3.6    Analytical  Procedures (for each analytical system)

         !M!   Instrument performance specifications;

         :x:   Instrument operating procedures;

         00   Data acquisition  system operation;

         :M:   Procedures when automatic quantitation algorithms are  overridden;

         •*•'   QC required parameters;

         ™-   Analytical run/injection logbooks; and

         :m-   Instrument error  and editing flag descriptions and  resulting
              corrective actions.

 6.3.7    Maintenance Activities  (for each analytical system)

          '•*'   Preventative  maintenance schedule and procedures,

          :*!   Corrective maintenance determinants  and procedures,  and

          '•^   Maintenance authorization.

 6.3.8    Analytical  Standards

          "=   Standard coding/identification  and inventory  system;

          ^   Standards  preparation  logbook(s);

          '•*"•   Standard preparation procedures;




 ILM05.2                               E-16

-------
                                                          Exhibit.E  — Section 6
                               Standard Operating Performance Standards  (Con't)

         ^   Procedures  for equivalency/traceability analyses  and
             documentation;

         •*•   Purity logbook  (primary  standards  and solvents) ;

         •*•   Storage,  replacement,  and labeling requirements;  and

         ^   QC and corrective  action measures.

6.3.9    Data Reduction  Procedures

         ^   Data processing  systems  operation;

         •*••   Outlier  identification methods;

         *:   Identification of  data requiring corrective action;  and

         •*•   Procedures  for format  and/or forms for each operation.

6.3.10   Documentation Policy/Procedures

         ^   Contractor/analyst's notebook policy,  including review policy;

         =*=   Complete Sample  Delivery Group (SDG)  File (CSF)  contents;

         :*:   Complete SDG  File  organization and assembly procedures,  including
             review policy; and

         ^   Document inventory procedures,  including review policy.

6.3.11   Data Validation/Self-Inspection  Procedures

         :*!   Data flow and chain-of-command for data review;

         ^   Procedures  for measuring precision and accuracy;

         •*'•   Evaluation  parameters  for identifying systematic  errors;

         ^   Procedures  to assure that hardcopy and electronic deliverables
             are complete  and compliant with the requirements  in  the Statement
             of Work  (SOW) Exhibits B and H;

         '•*'•   Procedures  to assure that hardcopy deliverables are  in agreement
             with their  comparable  electronic deliverables;

         ^   Demonstration of internal Quality  Assurance (QA)  inspection
             procedure (demonstrated  by supervisory sign-off on personal
             notebooks,  internal laboratory evaluation samples, etc.);

         'M:   Frequency and type of  internal audits (e.g.,  random,  quarterly,
             spot checks,  perceived trouble areas);

         ••*=   Demonstration of problem identification,  corrective  actions,  and
             resumption  of analytical processing.   Sequence  resulting from
             internal audit  (i.e.,  QA feedback);  and

         -*-•   Documentation of audit reports (internal and external),  response,
             corrective  action,  etc.
                                      E-17                               ILM05.2

-------
Exhibit E — Section 6
Standard Operating Performance Standards  (Con't)

6.3.12   Data Management  and  Handling

         •x:  Procedures  for controlling and estimating data entry errors;

         ^  Procedures  for reviewing  changes to data and deliverables and
             ensuring  traceability of  updates;

         "•  Lifecycle management procedures for testing, modifying,  and
             implementing changes to existing computing systems including
             hardware,  software,  and documentation or installing new systems;

         ^  Database  security,  backup, and archival procedures including
             recovery  from system failures;

         •*:  System maintenance  procedures and response time;

         =*•   Individual(s) responsible for system operation, maintenance,  data
              integrity,  and security;  and

         •x:   Specifications for  staff training procedures.

6.4    Updating  and Submitting SOP Requirements

6.4.1    The revised SOPs will become  the official SOPs under the contract and
         may be used during legal proceedings.  The Contractor shall maintain
          the complete  set of SOPs on file at the Contractor's facility for the
          term  of the contract.  Both the initial submission and the revised
          SOPs  shall be paginated consecutively in ascending order.  The
          revised SOPs  shall include:

          :*:   Changes resulting from (1) the Contractor's internal review of
              their procedures and  (2)  the Contractor's implementation of the
              requirements of the contract, and

          '-*1   Changes resulting from USEPA' s review of the laboratory
              evaluation sample data, bidder supplied documentation, and
              recommendations made during the pre-award on-site laboratory
              evaluation.

 6.4.1.1     The Contractor shall send a complete set of the latest version of
             SOPs or individually requested SOPs within 7 days of a request
             from an USEPA Regional CLP PO or the USEPA OERR AOC PM.  The
             request will designate the recipients.

 6.4.2     Subsequent Updates and Submissions.  During the term of the contract,
          the Contractor shall amend the SOPs when the following circumstances
          occur:

          :Ť   USEPA modifies  the technical requirements of the SOW or contract;

          •ť=   USEPA notifies  the Contractor of deficiencies  in the SOP
              documentation;

          :x:   USEPA notifies  the Contractor of deficiencies  resulting from
              USEPA's review  of the Contractor's performance;

          ^   The Contractor's procedures change;

          ^   The Contractor  identifies deficiencies  resulting from the
              internal review of the SOP documentation; or
 ILM05.2                               E-18

-------
                                                         Exhibit E — Section 6
                               Standard Operating Performance Standards  (Con't)

         ^   The Contractor identifies  deficiencies resulting from the
             internal review of their procedures.

6.4.2.1     Existing SOPs shall be amended or new SOPs shall be written within
            14 days of when the circumstances listed in Exnibit E, Section
            6.4, result in a discrepancy between what was previously described
            in the SOPs and what is presently occurring at the Contractor's
            facility.  All changes in the SOPs shall be clearly marked  (e.g.,
            a bar in the margin indicating where the change  is in the
            document, highlighting the  change by underlining the change, bold
            printing the change, or using a different print  font) and a copy
            is sent to the USEPA Regional CLP PO and Quality Assurance
            Technical Support  (QATS).   The amended/new SOPs  shall have the
            date on which the changes were implemented.

6.4.2.2     When existing SOPs are amended or new SOPs are written, the
            Contractor shall document the reasons for the changes and maintain
            the amended SOPs or new SOPs on file.  Documentation of the
            reasons for the changes shall be maintained on file with the
            amended SOPs or new SOPs.

6.4.2.3     Documentation of the reason(s) for changes to the SOPs shall also
            be submitted along with the SOPs.

6.5   Incentives/Sanctions

      The Contractor shall amend SOPs as specified within this section.  The
      .SOPs specify analytical procedures in greater detail than appear in
      Exhibit D.  Adherence to these requirements will ensure that the
      procedure is conducted in a standard, reliable, and reproducible process
     -•described in ILM05.2.  Failure to comply with the requirements specified
      herein may result in sanctions as described in the contract.
                                      E-19                               ILM05.2

-------
Exhibit E — Section 7
Contract Compliance Screening Performance Standards

7.0   CONTRACT COMPLIANCE SCREENING  (CCS) PERFORMANCE STANDARDS

7.1   Overview

7.1.1    CCS is one aspect of the Government's contractual right of inspection
         of analytical  data.  CCS examines  the Contractor's adherence to  the
         contract  requirements based on  the Sample Data Package delivered to
         USEPA.

7.1.2    CCS is performed by the Sample  Management Office  (SMO) under the
         direction of USEPA.  To assure  a uniform review,  a set of
         standardized procedures has been developed  to evaluate the Sample
         Data  Package submitted by a Contractor  against the technical and
         completeness requirements of  the contract.  USEPA reserves the right
         to add and/or  delete individual checks.

7.2   CCS Results

      CCS results  are distributed to the Contractor  and other  data recipients.
      The Contractor has  4 business  days to correct  deficiencies and shall
      send all corrections to the Regional  client and SMO.  CCS results are
      used in  conjunction with other information to  measure overall Contractor
      performance  and to  take appropriate actions to correct deficiencies in
      performance.

7.3   CCS Trend Report

      USEPA will periodically generate a CCS  trend report  which summarizes CCS
      results  over a  given period of time.  USEPA will send the CCS trend
      report  or discuss the  CCS  trend  report  during  an on-site laboratory
      evaluation.   In a detailed letter to  the USEPA Regional  Contract
      Laboratory  Program Project Officer (CLP PO) and USEPA Contracting
      Officer, the Contractor shall  address the  deficiencies and the
      subsequent  corrective  action  implemented by the Contractor to correct
      the  deficiencies  within 14 days  of receipt of  the report or  the on-site
      laboratory  evaluation.

 7.4   Incentives/Sanctions

 7.4.1     If  new  Standard Operating  Procedures (SOPs) are required  to be
          written,  or if existing SOPs  are  required  to be rewritten or amended
          because  of deficiencies and subsequent  corrective action  implemented
          by the Contractor,  the  Contractor  shall write/amend the SOPs per the
          requirements listed in  Exhibit E,  Section  6.

 7.4.2     The  Contractor shall correct  deficiencies  and resubmit the data
          within 4 business  days,  as specified within this  section.
          Resubmission and correction of the data will ensure that  the end user
          is reviewing contractually compliant data  described in  ILM05.2.
          Correct resubmission of the data may also  result  in a reduction  in
          overall sanctions.   Specific  details on incentives can be found  in
          the contract.   If the  Contractor  fails  to  adhere  to the requirements
          listed in this section, the Contractor  will be  in noncompliance  with
          the contract and may be subjected to sanctions  as described  in  the
          contract.
 'ILM05.2                               E-20

-------
                                                         Exhibit E —' Section 8
                                  Analytical  Performance  Standards  Requirements

8.0   ANALYTICAL PERFORMANCE STANDARDS REQUIREMENTS

8.1   Overview

      USEPA will not supply analytical reference standards either for direct
      analytical measurements or for the purpose of traceability.   All
      contract laboratories shall be required to prepare from materials or
      purchase from private chemical supply houses those standards necessary
      to successfully and accurately perform the analyses required in this
      protocol.

8.2   Preparation of Chemical Standards from the Neat High Purity Bulk
      Material

8.2.1    If the laboratory cannot obtain analytical reference standards, the
         laboratory may prepare their own chemical standards.  Laboratories
         shall obtain the highest purity possible when purchasing chemical
         standards; standards purchased at less than 97% purity shall be
         documented as to why a higher purity could not be obtained.

8.2.2    The chemical standards shall be kept at manufacturer recommended
         conditions when not being used in the preparation of standard
         solutions.  Proper storage of chemicals is essential in order to
         safeguard them from decomposition.

8.2.3    The Contractor shall be responsible for having analytical
         documentation proving the purity of each compound as stated.  Purity
         confirmation, when performed, shall use appropriate techniques.  Use
         of two or more independent methods is recommended.  The correction
         factor for impurity when weighing neat materials in the preparation
         of solution standards is:

         EQ. 1  Weight of Impure Compound

                 .  . .   _ .              ,   weight of pure compound
               weight of impure compound =	-^	
                                            (percent purity/100)
         where "weight of pure compound" is that required to prepare a
         specific volume of a solution standard of a specified concentration.

8.2.4    The Contractor is responsible for obtaining analytical documentation
         proving that all compounds used in the preparation of solution
         standards are correctly identified.

8.2.5    Logbooks shall be kept for all weighing and dilutions.  All
         subsequent dilutions from the primary standard and the calculations
         for determining their concentrations shall be recorded and verified
         by a second person.  All solution standards shall be refrigerated, if
         required, when not in use.  All solution standards shall be clearly
         labeled as to the identity of the analyte or analytes, the standard
         ID number of the solution, concentration, date prepared, solvent,
         expiration date of the solution, special storage requirements  (if
         any), and initials of the preparer.

8.3   Purchase of Chemical Standards Already in Solution

      Solutions of analytical reference standards can be purchased by
      Contractors provided the solutions meet the following criteria.
                                      E-21                               ILM05.2

-------
Exhibit E — Section 8
Analytical Performance Standards Requirements  (Con't)

8.3.1    Reference standards  shall be  accompanied by documentation  of  the
         purity confirmation  of  the  material  to  verify  the  integrity of the
         standard solutions.

8.3.2    The quality of reference standards purchased shall be  demonstrated
         statistically and analytically by a  method of  the  supplier's  choice.
         One way this can  be  demonstrated is  to  prepare and analyze three
         solutions: a high standard,  a low standard, and a  standard at the
         target concentration (see Sections  8.3.2.1 and 8.3.2.2").   The
         supplier must then demonstrate that  the analytical results for the
         high  standard and low standard are  consistent  with the difference in
         theoretical concentrations.   This is done  by the Student's t-test in
         Section 8.3.2.4.   If this is  achieved,  the supplier  must  then
         demonstrate that  the concentration of the  target standard lies midway
         between the concentrations  of the low and  high standards.   This is
         done  by the Student's t-test  in Section 8.3.2.5.  Thus,  the standard
         is  certified to be within 10% of the target concentration using the
         equations in Section 8.3.2.6.  If the procedure above  is  used, the
         supplier must document  that the following  have been  achieved.

8.3.2.1      Two solutions  of  identical concentration shall  be prepared
             independently  from neat  materials.  An  aliquot  of the first
             solution  shall be diluted to the intended  concentration (the
             "target  standard").   One aliquot is taken  from  the  second  solution
             and diluted  to a  concentration 10% greater  than the target
             standard.  This is called the "high standard".   One further
             aliquot  is  taken  from the second solution  and diluted to a
             concentration  10% less  than the target  standard.   This is  called
             the "low standard".

8.3.2.2      Six replicate  analyses of each standard (a  total  of 18 analyses)
             shall  be performed in the following sequence: low standard; target
             standard;  high standard; low standard;  target standard; high
             standard;  etc.

8.3.2.3      The mean  and variance of the six results for each solution shall
             be calculated:
             EQ. 2  Mean
                         MEAN =
                                         Y3 + Y4
           Y5 +
             EQ. 3  Variance

                           Y,2 +• Y,2 + Y,2 + Y 2 + Y 2 + Y 2  - 6 (MEAN)2
                VARIANCE = -i	?	1	*-	5	S	—
             The values Ylf  Y2,  Y3,  ...,  represent  the results of the  six
             analyses of each standard.  The means of the low, target, and high
             standards are designated M lt M2, and M3,  respectively.   The
             variances of the low, target, and high standards are designated
             Vj,  V2,  and V3, respectively.  Additionally, a pooled variance, V p,
             is calculated.
 ILM05.2
E-22

-------
                                                         Exhibit E — Section 8
                         Analytical  Performance Standards Requirements  (Con't)
            EQ. 4  Pooled Variance

                                   V           V
                                                3
                                                3
                                =  0-81    2   1.21
            If the square root of Vp is less than one percent of M2,  then
            M22/10,000 is to be used as the value of Vp in all subsequent
            calculations.

8.3.2.4     The test statistic shall be calculated:

            EQ. 5  Low and High Standard Test Statistic
                         TEST STATISTIC =
                                            M3     Ml
                                            1.1    0.9
                                              VP
            If the test statistic exceeds 2.13, then the supplier has failed
            to demonstrate a 20% difference between the high and low
            standards.  In such a case, the standards are not acceptable.

8.3.2.5     The test statistic shall be calculated:

            EQ. 6  Target Standard Test Statistic


                                             M, ^    f  M3
                     TEST STATISTIC =      >  -•-,    ť 2.2
            If the test statistic exceeds 2.13, the supplier has failed to
            demonstrate that the target standard concentration is midway
            between the high and low standards.  In such a case, the standards
            are not acceptable.

8.3.2.6     The 95% confidence intervals for the mean result of each standard
            shall be calculated:

            EQ. 7  Low Standard Interval

                                                       , v.
                  Interval for Low Standard = M, ą 2.13 | —Ł
                                                         6
            EQ. 8  Target Standard Interval

                 Interval for Target  Standard = M2 ą 2.13  -
                                      E-23                              ILM05.2

-------
Exhibit E — Section 8
Analytical Performance Standards Requirements  (Con't)

            EQ.  9   High  Standard Interval

                                                          n
                  Interval for High Standard = M3 ą 2.13  -j-
8.3.2.6.1      These  intervals  shall not overlap.   If overlap is observed,
               then the supplier  has  failed to demonstrate  the  ability to
               discriminate the 10% difference in  concentrations.   In such a
               case,  the standards are not acceptable.

8.3.2.6.2      In  any event,  the  Contractor is responsible  for  the  quality of
               the standards employed for  analyses under  this contract.

8.4   Requesting Standards  from the USEPA  Standards Repository

      Solutions of analytical reference materials  can be  ordered from the
      USEPA Chemical  Standards  Repository,  depending on availability.   The
      Contractor may  place  an order for standards  only after demonstrating
      that these standards  are  not available from  commercial vendors,  either
      in solution  or  as a neat  material.

8.5   Documentation of the  Verification and Preparation of  Chemical Standards

      It is the responsibility of the  Contractor to maintain the necessary
      documentation to show that  the  chemical standards it  has  used in the
      performance  of  Contract Laboratory Program (CLP) analysis conform to the
      requirements previously listed.

8.5.1     Weighing  logbooks,  calculations,  raw data, etc., whether produced by
          the Contractor or  purchased  from  chemical supply houses,  shall be
          maintained by the  Contractor and  may be subject  to review  during on-
          site inspection visits.   In  those cases where the  documentation is
          supportive of the  analytical results of data packages  sent to USEPA,
          such documentation is to be  kept  on file  by the  Contractor for a
          period  of one year.

8.5.2     Upon request by the USEPA Regional CLP Project Officer (CLP PO), the
          Contractor shall submit  their most recent previous year's
          documentation  (12  months)  for the verification and preparation of
          chemical  standards within  14 days of the  receipt of request to the
          designated recipients.

8.5.3     USEPA will periodically  generate  a report discussing deficiencies in
          the Contractor's documentation for the verification and preparation
          of chemical standards.  USEPA will send the report or  discuss the
          deficiencies during an on-site laboratory evaluation.   In  a detailed
          letter  to the USEPA Regional CLP  PO and CLP Quality Assurance
          Coordinator, the Contractor  shall address the deficiencies and the
          subsequent corrective action implemented by the  Contractor to correct
          the deficiencies within 14  days  of receipt of the  report  or the on-
          site laboratory evaluation.

8.5.4     If new  Standard Operating  Procedures (SOPs) are  required  to be
          written,  or if existing SOPs are  required to be  rewritten  or amended
          because of deficiencies and  subsequent corrective  action  implemented
          by the  Contractor, the Contractor shall write/amend the  SOPs per the
          requirements listed in Exhibit E, Section 6.
 ILM05.2                               E-24

-------
                                                         Exhibit E — Section 8
                         Analytical  Performance Standards Requirements  (Con't)

8.6   Incentives/Sanctions

      The Contractor shall obtain the highest purity possible when purchasing
      chemical standards specified within this section.  The use of high
      purity standards will ensure a more accurate identification and
      quantitation of analytes described in the ILM05.2 Statement of Work
      (SOW).  Failure to meet the requirements set forth in this section may
      result in sanctions as described in the contract.
                                      E-25                              ILM05.2

-------
Exhibit E — Section 9
Data Package Monitoring Audits

9.0   DATA PACKAGE MONITORING AUDITS

9.1   Overview

      Data package audits are performed by USEPA for program overview and
      specific Regional concerns.  Standardized procedures have been
      established to assure uniformity of the auditing process.  Data packages
      are periodically selected  from  recently received Cases.  They are
      evaluated for the technical quality of hardcopy raw data, Quality
      Assurance (QA), and adherence to contractual requirements.  This
      function provides external monitoring of program Quality Control  (QC)
      requirements.  Data package audits are used to assess the technical
      quality of the data and evaluate overall laboratory performance.  Audits
      provide USEPA with an in-depth  inspection and evaluation of the Case
      data package with regard to achieving QA/QC acceptability.  A thorough
      review of the raw data is  completed including: all instrument readouts
      used for the sample results, instrument printouts, and other
      documentation for deviations from the contractual requirements, a check
      for transcription and calculation errors, a review of the qualifications
      of the laboratory personnel involved with the Case, and a review of  the
      latest version of all Standard  Operating Procedures  (SOPs) on file.

9.2   Responding to the Data Package  Audit Report

9.2.1    After completion of the data package audit, USEPA will send a copy of
         the data package audit  report  to  the Contractor or discuss the data
         package audit  report  on an on-site  laboratory  evaluation.  In a
         detailed letter  to the  USEPA Regional Contract Laboratory  Program
         Project Officer  (CLP  PO) and the  USEPA designated recipient, the
         Contractor  shall discuss the corrective actions implemented to
         resolve the deficiencies listed  in  the data package  audit  report
         within 14  days  of  receipt of the  report.

9.2.2    If new SOPs are  required to  be written, or if  existing SOPs are
         required to be  rewritten or  amended because of deficiencies and
         subsequent  corrective action implemented by the Contractor, the
         Contractor  shall write/amend the  SOPs per  the  requirements listed in
         Exhibit E,  Section  6.

9.3    Incentives/Sanctions

       The Contractor  shall  discuss the  corrective actions  implemented to
       resolve  the deficiencies listed in  the data package  audit  report within
       14 days  of receipt  of  the  comments  from USEPA, as specified within  this
       section.  The  data  package audits ensure that the policies and
       procedures identified in this  Statement of Work  (SOW) meet the
       requirements  of  this  contract.   If  the Contractor fails  to adhere to the
       requirements  listed in  this  section,  the Contractor  will be  in
       noncompliance  with  the  contract and may be subjected to  sanctions as
       described in  the contract.
 ILM05.2                               E-26

-------
                                                        Exhibit E — Section 10
                                                Regional Data Review Monitoring

10.0  REGIONAL DATA REVIEW MONITORING

10.1  Overview

      Contractor data are generated to meet the specific needs of USEPA
      Regions.  In order to verify the usability of data for the intended
      purpose, each Region reviews data from the perspective of the end user,
      based on functional guidelines for data review which have been developed
      jointly by the Regions and the USEPA OERR Analytical Operations/Data
      Quality Center (AOC).  Each Region uses these guidelines as the basis
      for data evaluation.  Individual Regions may augment the basic guideline
      review process with additional review based on Region-specific or site-
      specific concerns.  Regional reviews, like the sites under
      investigation, vary based on the nature of the problem under
      investigation and the Regional response appropriate to the specific
      circumstances.

10.1.1   Regional data reviews, relating usability of the data to a specific
         site, are part of the collective assessment process.  They complement
         the review done at the Sample Management Office (SMO),< which is
         designed to identify contractual discrepancies, and the review done
         by the USEPA OERR AOC, which is designed to evaluate Contractor and
         method performance.
                                      E-27
ILM05.2

-------
Exhibit E — Section 11
QA Proficiency Monitoring

11.0  QUALITY ASSURANCE  (QA) PROFICIENCY MONITORING

      As a means of measuring and evaluating both the Contractor's and the
      method's analytical performance, the Contractor shall participate in
      USEPA's Proficiency Testing Program.  USEPA's Proficiency Testing
      Program involves the  analysis of Case specific Performance Evaluation
       (PE) samples and Quarterly Blind  (QB) Audits.  The Contractor's
      analytical PE samples and QB results will be used by USEPA to assess and
      verify the Contractor's continuing ability to produce acceptable
      analytical data in accordance with the contractual requirements.  The
      Contractor shall receive a passing score of 75% to be in compliance with
      the contract.

11.1  Performance Evaluation  (PE) Samples

11.1.1   The PE sample(s) may be scheduled with the Contractor as frequently
          as on a Sample  Delivery Group  (SDG)-by-SDG basis.  The  PE  samples may
          be sent either  by  the Regional  client or the USEPA OERR Analytical
          Operations/Data Quality Center  (AOC).  PE samples assist USEPA in
          monitoring Contractor performance.

11.1.2    PE samples will be provided  as  either single-blinds  (recognizable as
          a  PE sample but of unknown composition), or as  double-blinds  (not
          recognizable  as a  PE sample  and of unknown composition).   The
          Contractor will not  be  informed of either the analytes/parameters or
          the concentrations in the PE samples.

11.1.3    The Contractor  may receive the  PE samples as either  full volume
          samples or ampulated/bottled concentrates from  USEPA or a  designated
          USEPA Contractor.  The  PE samples shall  come with instructions
          concerning the  unique preparation procedures, if any,  required to
          reconstitute  the  PE  samples  (i.e., the required dilution  of  the  PE
          sample  concentrate)-.  PE  samples  are  to  be digested  and analyzed with
          the rest  of  the routine  samples in the SDG.  The Contractor  shall
          prepare and  analyze  the  PE  sample using  the procedure described  in
          the sample preparation  and method analysis  sections  of Exhibit D.
          All contract  required Quality Control  (QC)  shall be  met.   The  PE
          sample  results  are to be  submitted in  the SDG deliverable  package per
          normal  reporting  procedures  detailed  in  Exhibit B.

11.1.4    In addition  to  PE  sample preparation and analysis, the  Contractor
          shall be  responsible for  correctly identifying  and  quantitating  the
          analytes  included in each  PE sample.   When  PE  sample results  are
          received  by  USEPA, the  PE  sample  results will be  evaluated for
          correct analytical identification and quantitation.   The  PE  sample
          evaluation will be provided to the Contractor via  coded evaluation
          sheets, by analyte.   USEPA  will notify the  Contractor of  unacceptable
          performance.

 11.2  Quarterly  Blind (QB)  Audits

 11.2.1    A QB Audit is a unique  analytical Case containing  only PE  samples
          (i.e.,  referred to as  QB samples).   The  QB  samples  will be scheduled
          by the  USEPA OERR AOC  through the Sample Management  Office (SMO).   QB
          samples assist USEPA in monitoring  Contractor  performance.

 11.2.2    QB samples will be provided as single-blinds  (recognizable as  a  PE
          sample  but of unknown  composition).   The Contractor  will  not be
          informed  of  either the  analytes or  the concentrations in  the PE
          samples.
 ILM05.2                               E-28

-------
                                                        Exhibit E -- Section 11
                                              QA Proficiency Monitoring (Con't)

11.2.3   The Contractor may receive the QB samples as either full volume
         samples or ampulated/bottled concentrates from USEPA or a designated
         USEPA Contractor.  The QB samples shall come with instructions
         concerning the unique preparation procedures, if any, required to
         reconstitute the QB samples  (i.e., the required dilution of the QB
         sample concentrate).  The Contractor shall prepare and analyze the QB
         samples using the procedure described in the sample preparation and
         method analysis sections of Exhibit D.  All contract required QC
         shall be met, including spike and duplicate analyses.  The QB sample
         results are to be submitted in the SDG deliverable package per normal
         reporting procedures detailed in Exhibit B.

11.2.4   In addition to QB sample preparation and analysis,  the Contractor
         shall be responsible for correctly identifying and quantitating the
         analytes included in each QB sample.  When QB sample results are
         received by USEPA, the QB sample results will be scored for correct
         analytical identification and quantitation.  The QB sample scoring
         will be provided to the Contractor via coded evaluation sheets, by
         analyte.  USEPA will notify the Contractor of unacceptable
         performance.  The Contractor's QB sample performance will be assessed
         into one of the following three categories:
                                      i
11.2.4.1    Acceptable, No Response Required: Score greater than or equal to
            90%.  The data meets most or all of the scoring criteria.  No
            response is required.

11.2.4.2    Acceptable, Response Explaining Deficiencies Required: Score
            greater than or equal to 75%, but less than 90%.  Deficiencies
            exist in the Contractor's performance.  Corrective action response
            required.

11.2.4.3    Unacceptable Performance, Response Explaining Deficiencies
            Required: Score less than 75%.  Corrective action response
            required.

11.2.5   In the case of Section 11.2.4.2 or 11.2.4.3, the Contractor shall
         describe the deficiency(ies) and the action (s) taken in a corrective
         action letter to the USEPA Contracting Officer, USEPA Regional
         Contract Laboratory Program Project Officer  (CLP PO), and CLP Quality
         Assurance  (QA) Coordinator within 14 days of receipt of notification
         from USEPA.

11.2.6   In the case of Section 11.2.,4.2 or 11.2.4.3, if new Standard
         Operating Procedures  (SOPs) are required to be written, or if
         existing SOPs are required to be rewritten or amended because of
         deficiencies and subsequent corrective action implemented by the
         Contractor, the Contractor shall write/amend the SOPs per the
         requirements listed in Exhibit E, Section 6.

11.2.7   The Contractor shall be notified by the USEPA Contracting Officer
         concerning agreement or disagreement with the proposed remedy for
         unacceptable performance.

11.2.8   A Remedial QB Audit is a unique analytical Case containing only QB
         samples.  A Remedial QB Audit may be scheduled by the USEPA OERR AOC
         with the Contractor(s) for any of the following reasons: unacceptable
         PE sample performance, unacceptable QB sample performance, and/or
         major change in the laboratory  (e.g., relocation, new owner, or high
         turn-over of key personnel).  Sections 11.2.2 through 11.2.7 apply to
         the Remedial QB Audit process.
                                      E-29                               ILM05.2

-------
Exhibit E -- Section 11
QA Proficiency Monitoring  (Con't)

11.3  Incentives/Sanctions

      The Contractor shall analyze PE and QB samples with acceptable
      analytical results in accordance with the contractual requirements as
      described in this section.  If the Contractor fails to adhere to the
      requirements listed  in this section, the Contractor will be in
      noncompliance with the contract and may be subjected to sanctions as
      described in the contract.
 ILM05.2                              E-30

-------
                                                        Exhibit E — Section 12
                                   On-Site Laboratory QA Monitoring Evaluations

12.0  ON-SITE LABORATORY QUALITY ASSURANCE  (QA) MONITORING EVALUATIONS

12.1  Overview

      The USEPA Regional Contract Laboratory Program Project Officer  (CLP PO)
      or the USEPA Contracting Officer's authorized representative will
      conduct an on-site laboratory evaluation.  On-site laboratory
      evaluations are carried out to monitor the Contractor's ability to meet
      selected terms and conditions specified in the contract.  The evaluation
      process incorporates two separate categories: Quality Assurance  (QA)
      Evaluation and Evidentiary Audit.

12.2  Quality Assurance On-Site Evaluation

      QA evaluators inspect the Contractor's facilities to verify the adequacy
      and maintenance of instrumentation, the continuity, experience and
      education of personnel, and the acceptable performance of analytical and
      Quality Control  (QC) procedures for adherence to the contract
      requirements.

12.2.1   The Contractor shall expect that items to be monitored will include,
         but are not limited to, the following:

         ^   Size, cleanliness, and organization of the facility;

         'x:   Quantity, age, availability, scheduled maintenance, and
             performance of instrumentation;

         ="   Availability, appropriateness, and utilization of the Quality
             Assurance Management Plan  (QAP) and Standard Operating Procedures
              (SOPs);

         <*•   Staff qualifications, experience, and personnel training
             programs;

         =*=   Analysis of Performance Evaluation  (PE) sample (s) ;

         •*:   Reagents, standards, and sample storage facilities;

         "=   Standard preparation logbooks  and raw data;

         ^   Bench sheets and analytical'  logbook maintenance and review; and

         :i:   Review of the Contractor' s sample analysis/data package
             inspection/data management procedures.

12.2.2   Prior to an on-site evaluation, various documentation pertaining to
         performance of the specific Contractor is integrated into a profile
         package for discussion during the  evaluation.  Items that may be
         included are: previous on-site reports; Quarterly Blind (QB) and/or
         PE sample scores results; Regional review of data; Contractor
         performance information provided by the Region; data audit reports;
         results of Contract Compliance Screening  (CCS); and data trend
         reports.

12.3  Evidentiary Audit

      Evidence auditors conduct an on-site  laboratory evaluation to determine
      if laboratory policies and procedures are in place to satisfy evidence
      handling requirements as stated in Exhibit F.  The evidence audit
      comprises a procedural audit, an audit of written SOPs, and an audit of
      analytical project file documentation.


                                      E-31                               ILM05.2

-------
Exhibit E — Section 12
On-Site Laboratory QA Monitoring Evaluations  (Con't)

12.3.1   Procedural Audit.  The Contractor shall perform analysis of PE
         sample(s) in the presence of the USEPA designated team  during the
         procedural audit.  The procedural audit will be comprised of
         everything from 'sample receipt to data package assembly and
         completion.  This  includes the review and  examination of actual  SOPs
         and accompanying documentation for the following laboratory
         operations: sample receiving, sample storage, sample identification,
         sample security, sample tracking  (from receipt to completion of
         analysis), analytical project file organization and assembly, and
         proper disposal of samples and cogenerated wastes.

12.3.2   Written SOPs Audit.  The written SOPs audit consists of review and
         examination of the written SOPs to determine if they are accurate and
         complete  for the following laboratory operations: sample receiving,
         sample storage, sample identification, sample security, sample
         tracking  (from receipt to completion of analysis), and  analytical
         project  file organization and assembly.

12.3.3   Analytical Project File Evidence Audit.  The analytical project  file
         evidence  audit consists of review and examination of the analytical
         project  file documentation.  The auditors  review the files  to
         determine:

         •*•   The  accuracy  of  the document  inventory;

         "M:   The  completeness of the  file;

         '•*>   The  adequacy  and accuracy of  the document numbering system;

         "•   Traceability  of  sample activity;

         '•'•   Identification of activity  recorded  on the  documents;  and

         •*•   Error correction methods.

12.4   Discussion  of the On-Site Team's Findings

       The  QA and  evidentiary  auditors discuss their findings  with the  USEPA
       Regional CLP PO prior to debriefing  the Contractor.   During the
       debriefing,  the  auditors present their  findings and  recommendations for
       corrective  actions  necessary  to the  Contractor personnel.  A  report
       which discusses  deficiencies  found during the on-site  audit will be sent
       to the Contractor  to provide  further  clarification of  findings.   In a
       detailed letter  to  the  USEPA  Regional CLP PO  and CLP  Quality  Assurance
       Coordinator, the Contractor  shall  discuss the deficiencies and the
       subsequent  corrective actions implemented by  the Contractor to resolve
       the  deficiencies within 14 days of receipt  of report  or the on-site
       laboratory  evaluation.

 12.4.1  If new SOPs are  required  to  be  written,  or if existing  SOPs are
          required to  be  rewritten  or  amended  because of  the deficiencies  and
          the subsequent  corrective  action  implemented by the Contractor,  the
          Contractor shall write/amend the  SOPs  per  the  requirements listed in
          Exhibit  E,  Section 6.

 12.5   Incentives/Sanctions

       The  Contractor  shall submit  to  on-site  evaluations,  as  specified within
       this section.   The on-site evaluations  ensure that the policies and
       procedures  identified in this Statement of  Work (SOW)  meet the
       requirements of  this contract.   If the  Contractor  fails to adhere  to the
       requirements listed in this  section,  the  Contractor  will  be  in


 ILM05.2                              E-32

-------
                                                  Exhibit E — Sections 12 & 13
                                           Electronic Data QA Monitoring Audits

      noncompliance with the contract and may be subjected to sanctions as
      described in the contract.

13.0  ELECTRONIC DATA QUALITY ASSURANCE  (QA) MONITORING AUDITS

13.1  Overview

      Periodically, USEPA requests the instrument electronic data from
      Contractors for a specific Case in order to accomplish electronic data
      audits.  Generally, electronic data submissions and audits are requested
      for the following reasons.

      ™   Program overview;

      •*'   Indication of data quality problems;

      1X3   Support for on-site audits; and

      ="   Specific Regional requests.

13.1.1   Depending upon the reason for an audit, the instrument electronic
         data from a recent Case, a specific Case, or a laboratory evaluation
         sample may be requested.  Electronic data audits provide a mechanism
         to assess adherence to contractual requirements and to ensure the
         consistency of data reported on the hardcopy/electronic deliverables
         with that generated on analytical  instruments.  This function
         provides external monitoring of Program Quality Control  (QC)
         requirements and checks adherence  of the Contractor to internal
         Quality Assurance  (QA) procedures.  In addition, electronic data
         audits enable USEPA to evaluate the utility, precision, and accuracy
         of the analytical methods.

13.1.2   The Contractor shall store all raw and processed electronic
         analytical data in the appropriate instrument manufacturer's format,
         uncompressed, and with no security codes.  The data shall include all
         necessary data files for a complete reconstruction of the previously
         submitted hardcopy and electronic  deliverable data package.  All
         associated raw data files in the instrument manufacturer proprietary
         software format must be submitted  if those files contain data or
         instrumental parameters regarding  any analysis and or correction
         applied to an instrument or analytical result.  This instrument
         electronic data shall include data for all samples and all QC
         samples, including but not limited to: blanks, matrix spikes, post-
         digestion spikes, analytical spikes, duplicates, serial dilutions,
         Laboratory Control Samples  (LCSs), Contract Required Quantitation
         Limits  (CRQL) Check Standards  (CRIs), Interference Check Samples
          (ICSs), tunes, initial calibrations and verifications, and Continuing
         Calibration Verifications  (CCVs).  In addition, the Contractor shall
         supply raw data for the Method Detection Limit  (MDL) studies and
         Linear Range Analyses  (LRS) which  are used to set the MDL and LRV
         values for the year/quarter in which the Sample Delivery Group  (SDG)
         was analyzed.  The Contractor shall maintain a reference logbook of
         data files of EPA sample number, calibration data, standards, blanks,
         spikes, and duplicates.  The logbook shall include EPA sample
         numbers, identified by Case and SDG.

13.1.3   The Contractor is required to retain the instrument electronic data
         for three years after submission of the reconciled Complete SDG File.
         Electronic media shipped to the USEPA designated recipient must be
         fully usable by the recipient.  Diskettes must be 3.5 inch, high
         density, 1.44 MB MS/DOS formatted  and tapes must be either 4 mm or 8
         mm. Alternative means for delivery of electronic data may be utilized


                                      E-33                .               ILM05.2

-------
Exhibit E — Section 13
Electronic Data QA Monitoring Audits  (Con't)

         by the Contractor  upon prior written  approval by USEPA.   When
         submitting electronic instrument  data to USEPA,  the following
         materials shall  be delivered in response to the  request.

13.1.3.1    All as'sociated  raw data  files  for  all analytical samples  and all
            QC  samples.   For example,  files for ICP should include raw
            intensities and mercury  and cyanide files should include  raw
            absorbances or  integrated areas.

13.1.3.2    All processed data files and quantitation output files associated
            with  the  raw  data files  described  in Section  13.1.3.1.

13.1.3.3    All associated  identification  and  calculation files used  to
            generate  the  data submitted in the data package.  This includes,
            but is not limited to,  result  files, acquisition files,
            calibration files, and method  files.

13.1.3.4    All Contractor-generated Inductively Coupled  Plasma -  Atomic
            Emission  Spectrophotometer (ICP-AES)/ICP - Mass Spectrophotometer
             (ICP-MS)  interference correction files must be submitted.

13.1.3.5    A  copy of the Contractor's reference logbook  relating  data files
            to EPA  sample number, calibration  data, standards, blanks, spikes,
            and duplicates.  The  logbook  shall include EPA sample numbers and
            laboratory file identifiers for all samples,  blanks, and
            standards, identified by Case  and  SDG.

13.1.3.6    A  printout of the directory of all files in each directory,
            including all subdirectories  and the files contained therein.

13.1.3.7    A  copy  (hardcopy)  of  the completed Sample Data Package.

13.1.3.8    A  statement attesting to the completeness of  the electronic
             instrument data submission, signed and dated by the Contractor's
             laboratory manager.   The Contractor shall also provide a statement
             attesting that the data reported have not been altered in any way.
             These statements shall be part of  a Cover Sheet that includes the
             following information relevant to  the data submission:

             5X3   Contractor name;

             =*•   Date of  submission;

             !ť:   Case number;

             ••ť   SDG number;

             :x:   Instrument make  and model number  for  each instrument;

             :x:   Instrument operating.software name and version number;

             •-*1   Data  software name  and version  used  for  acquisition, re-
                 quantitation, and hardcopy/report generation;

             :l-   Data  system computer;

             •*•   System operating software;

             ^   Data  system network;

             'x:   Data  backup software;
 ILM05.2                               E-34

-------
                                                        Exhibit E — Section 13
                                   Electronic Data QA Monitoring Audits (Con't)

            •*=   Data backup hardware;

            :m:   Media type and volume of data  (in MB) backed up; and

            '•^   Names and telephone numbers  of two Contractor contacts  for
                further information regarding the submission.

13.2  Submission of the Instrument Electronic Data

      Upon request of the USEPA Regional Contract Laboratory Program Project
      Officer (CLP PO), the Contractor shall send the required instrument
      electronic data and all necessary documentation to the USEPA designated
      recipient [e.g., Quality Assurance Technical Support  (QATS)]  within 7
      days of notification.

      NOTE: The instrument electronic data shall be shipped according to the
      procedures in Exhibit F.

13.3  Responding to the Electronic Data Audit Report

      After completion of the electronic data audit, USEPA will send a copy of
      the electronic data audit report to the Contractor or may discuss the
      electronic data audit report at an on-site laboratory evaluation.  In a
      detailed letter to the USEPA Regional  CLP PO, the Contractor shall
      discuss the corrective actions implemented to resolve the deficiencies
      listed in the electronic data audit report within 14 days of receipt of
      the report or the on-site laboratory evaluation.

13.3.1  If new Standard Operating  Procedures  (SOPs)  are  required to be  written
        or SOPs  are required to be amended because of the  deficiencies  and the
      - subsequent corrective action implemented by the  Contractor, the
        Contractor shall write/amend and submit the SOPs per the requirements
      .'  listed  in  Exhibit E,  Section 6.

13.4  Incentives/Sanctions

      /The Contractor shall submit to electronic data audits and adhere to the
      requirements specified in this section.  Resubmission and correction of
      electronic data will ensure that the end user is reviewing contractually
      compliant data described in the ILM05.2 contract.  If the Contractor
      fails to adhere to the requirements listed in this section, the
      Contractor will be in noncompliance with the contract and may be
      subjected to sanctions as described in the contract.
                                      E-35                               ILM05.2

-------
Exhibit E — Section 14
Data Management Performance Requirements

14.0  DATA MANAGEMENT PERFORMANCE REQUIREMENTS

14.1  Overview

14.1.1   Data management procedures are defined as procedures specifying the
         acquisition or entry,  update, correction, deletion, storage,  and
         security  of computer  readable data  and files.   These procedures shall
         be  in written form  and contain a  clear definition  for  all  databases
         and files used to generate or resubmit deliverables.   Key  areas of
         concern include:  system organization  (including personnel  and
         security), documentation operations,  traceability, and Quality
         Control  (QC).

14.1.2   Data manually entered from hardcopy shall be subject to QC checks and
         the error rates estimated.   Systems should prevent entry of  incorrect
         or  out-of-range data  and alert data entry personnel of errors.  In
         addition,  data entry  error rates  shall be estimated and recorded on  a
         monthly basis by  re-entering a statistical sample  of the data entered
         and calculating discrepancy  rates by  data element.

14.2  Documenting  Data Changes

      The  record of changes  in the form of corrections  and  updates  to data
      originally generated,  submitted, and/or  resubmitted shall be  documented
      to allow traceability  of updates.  Documentation  shall include  the
      following for each change.

      -JU   Justification  or rationale  for the change.

      1X3   Initials of  the  person making the  change (s).   Data  changes  shall be
           implemented  and  reviewed by a person or group independent of the
           source  generating the deliverable.

      ^   Documentation  of changes  shall be  retained  according  to  the schedule
           of the  original  deliverable.

      !*'   Resubmitted  diskettes or  other  deliverables shall be  re-inspected  as
           a part  of the  laboratory's  internal  inspection process prior to
           resubmission.   The entire  deliverable,  not  just the  changes, shall
           be inspected.

      '•*>   The  Laboratory Manager shall  approve changes  to originally submitted
           deliverables.

      '•*>   Documentation of data changes may  be requested by laboratory
           auditors.

 14.3  Lifecycle Management Procedures

       Lifecycle management procedures shall  be applied  to computer  software
       systems  developed by the Contractor  to be used  to generate and edit
       contract deliverables.  Such  systems shall  be thoroughly  tested and
       documented  prior to utilization.

 14.3.1    A software test and acceptance plan including  test requirements,  test
          results  and acceptance criteria  shall be developed,  followed, and
          available in written form.

 14.3.2    System changes  shall  not be  made  directly  to production systems
          generating deliverables.   Changes shall be made first  to  a
          development system and tested prior to implementation.
 ILM05.2                               E-36

-------
                                                        Exhibit E  —  Section 14
                               Data Management Performance Requirements (Con't)

14.3.3   Each version of the production system will be given an identification
         number, date of installation, and date of last operation  and will be
         archived.

14.3.4   System and operations documentation shall be developed and maintained
         for each system.  Documentation shall include a  user's manual and an
         operations and maintenance manual.

14.3.5   This documentation shall be available for on-site review  and/or upon
         written request by the USEPA Regional Contract Laboratory Program
         Project Officer  (CLP PO) or the USEPA OERR Analytical Operations/Data
         Quality Center  (AOC) Inorganic Program Manager  (AOC PM).

14.4  Personnel Responsibilities

      Individual(s) responsible for the following functions shall  be
      identified.

      ="   System operation and maintenance including documentation and
          training.

      ="   Database integrity, including data entry, data  updating  and QC.

      ™   Data and system security, backup and archiving.
                                      E-37                               ILM05.2

-------
Exhibit E — Section 15
Tables

15.0  TABLES

TABLE 1. Contract Laboratory Program Quality Assurance Monitoring Plan
Exhibit A:
Summary of
Requirements
Exhibit B: Reporting
and Deliverables
Requirements
Exhibit C: Inorganic
Target Analyte List
with Contract
Required
Quantitation Limits
Exhibit D:
Analytical Methods



Exhibit E: Contract
Laboratory Program
Quality Assurance
Monitoring Plan
Summary of Program
Requirements
Reporting and
Deliverable
Requirements
Target Analyte List
with Contract
Required
Quantitation Limits
ICP-AES requirements
are outlined in
Exhibit D, Part A,
Sections 1.0 through
8.0, 14.0, and 15.0.
ICP-MS requirements
are outlined in
Exhibit D, Part B,
Sections 1 . 0 through
8.0, 14.0, and 15.0.
Mercury requirements
are outlined in
Exhibit D, Part C,
Sections 1 . 0 through
8.0, 14.0 and 15.0.
Cyanide requirements
are outlined in
Exhibit D, Part D,
Sections 1 . 0 through
8.0, 14.0, and 15.0.
General QA/QC
Requirements
Performance
standards are
summarized in
Exhibit A, Sections
1 . 0 through 4.0.
Performance
standards are
outlined in Exhibit
B, Sections 1.. 0
through 4.0.
Performance
standards are
outlined in Exhibit
C, Section 1.0.
Performance
standards are
outlined in Exhibit
D, Part A, Sections
9.0 through 11.0.
Performance
standards are
outlined in Exhibit
D, Part B, Sections
9.0 through 11.0.
Performance
standards are
outlined in Exhibit
D, Part C, Sections
9.0 through 11.0.
Performance
standards are
outlined in Exhibit
D, Part D, Sections
9.0 through 11.0.
As outlined in
Exhibit D, Quality
Control sections.
QA monitoring plan
is outlined in
Exhibit E.
CCS in Exhibit E,
Section 7.0, and
CADRE will be used
to monitor reporting
electronic
deliverables.
QA monitoring plan
is outlined in
Exhibit E.
QA monitoring plan
is outlined in
Exhibit D, Part A,
Section 12.0, and
Exhibit E.
QA monitoring plan
is outlined in
Exhibit D, Part B,
Section 12.0, and
Exhibit E.
QA monitoring plan
is outlined in
Exhibit D, Part C,
Section 12.0, and
Exhibit E.
QA monitoring plan
is outlined in
Exhibit D, Part D,
Section 12.0, and
Exhibit E.
QA Management Plan
is outlined in
Exhibit E, Section
5.0.
 ILM05.2
E-38

-------
                                                              Exhibit E — Section 15
                                                                       Tables  (Con't)

     TABLE 1. Contract Laboratory  Program Quality Assurance  Monitoring Plan (Con't)
Exhibit E:  Contract
Laboratory Program
Quality Assurance
Monitoring Plan
(Con't)
Quality Assurance
Management Plan
As outlined in
Exhibit E, Sections
5.1.1 and 5.1.2, a
written QA
Management Plan
shall be used to
ensure acceptable
data production of
known and documented
quality.
USEPA will review
and approve the QA
Management Plan.
                       Standard  Operating
                       Procedures
                       Performance
                       standards are
                       outlined in Exhibit
                       E, Sections 6.0
                       through 6.4, and
                       must be performed as
                       stated.
                       SOPs will be
                       reviewed by USEPA
                       during Pre-Award,
                       on-site audits,
                       after modifications
                       are made and
                       randomly, as deemed
                       appropriate.
                        Contract  Compliance
                        Screening
                        Performance
                        standards are
                        outlined in Section
                        E.2 of the ILM05.2
                        IFB and must be
                        performed as stated.
                       The sample data
                       package will be
                       evaluated against
                       the technical and
                       completeness
                       requirements of the
                       contract.
                       Analytical  Standards
                       Performance
                        standards  are
                        outlined in Exhibit
                        E,  Sections 8.0
                        through 8.5, and
                        must be performed  as
                        stated.
                       Randomly, USEPA will
                       review analytical
                       standards
                       verification and
                       preparation
                       documentation, as
                       deemed appropriate.
                        Data Package Audits
                       Performance
                       standards  are
                       outlined in Exhibit
                       E,  Sections 9.0
                       through 9.2.
                       Data package audits
                       are performed by
                       USEPA to evaluate
                       technical quality of
                       the hardcopy raw
                       data, QA, and
                       adherence to
                       contractual
                       requirements.
                        Regional Data Review
                       Analytical data is
                        reviewed by each
                        Region  from the
                        perspective of the
                        end user to
                        determine  the
                        usability  of the
                        data, as outlined in
                        Exhibit E, Section
                        10.0.
                       Regional validation
                       and/or CADRE reports
                       are generated for
                       all data packages.
                                            E-39
                                                       ILM05.2

-------
Exhibit E — Section 15
Tables  (Con't)

TABLE 1. Contract Laboratory Program Quality Assurance Monitoring Plan  (Con't)
Exhibit E: Contract
Laboratory Program
Quality Assurance
Monitoring Plan
(Con't)



Exhibit F: Chain-of-
Custody, Document
Control and Written
Standard Operating
Procedures

Exhibit G: Glossary
of Terms
Proficiency Testing
On-Site Laboratory
Evaluations
Electronic Data
Audits
Data Management
Standard Operating
Procedures
Written Standard
Operating Procedures
Glossary of Terms
Performance
standards are
outlined in Exhibit
E, Sections 11.0
through 11.2, and
must be performed as
stated.
Performance
standards are
outlined in Exhibit
E, Sections 12.0
through 12.4.
Performance
standards are
•outlined in Exhibit
E, Sections 13.0
through 13.3.
Performance
standards are
outlined in Exhibit
E, Sections 14.0
through 14.4, and
must be performed as
stated.
Performance
standards are
outlined in Exhibit
F, Sections 2.0
through 2.7.
Performance
standards are
outlined in Exhibit
F, Sections 3.0
through 3.7.
Contractors shall
adhere to
interpretation of
SOW terms as defined
within Exhibit G.
Acceptable QB scores
will assist in
monitoring
contractor
performance as
defined in Exhibit
E, Sections 11.2.4.1
through 11.2.4.3,
and 11.2.8.
USEPA will evaluate
the results from
quality assurance
and evidentiary on-
site audits as
defined in Exhibit
E, Sections 12.2.1
through 12.3.3, to
assist in monitoring
the contractor.
CCS in Exhibit E,
Section 7.0, will be
used to monitor
electronic
deliverables .
USEPA will monitor
data management
practices during
quality assurance
and evidentiary on-
site audits.
SOPs will be
reviewed by USEPA
during Pre-Award,
on-site audits,
after modifications
are made, and
randomly as deemed
appropriate .
SOPs will be
reviewed by USEPA
during Pre-Award,
on-site audits,
after modifications
are made, and
randomly as deemed
appropriate.
N/A
 ILM05.2
E-40

-------
                                                              Exhibit E —  Section  15
                                                                       Tables (Con't)

     TABLE 1. Contract Laboratory  Program Quality Assurance Monitoring Plan  (Con't)
Exhibit H:
Data Dictionary and
Format for Data
Deliverables in
Computer-Readable
Format
           Data Dictionary and
           Format
                       Performance
                       standards are
                       outlined in Exhibit
                       H and Appendix A.
                       CCS in Exhibit E,
                       Section 7.0,  will be
                       used to monitor
                       electronic
                       deliverables.
Appendix B:
Analysis
Modified
GFAA requirements
are outlined in
Appendix B, Sections
1.0 through 8.0,
14.0, and 15.0.
Performance
standards are
outlined in Appendix
B, Sections 9.0
through 11.0.
QA monitoring plan
is outlined in
Appendix B, Section
12.0, and Exhibit E.
                                           E-41
                                                                  ILM05.2

-------
THIS PAGE INTENTIONALLY LEFT BLANK

-------
                EXHIBIT  F

    CHAIN-OF-CUSTODY,  DOCUMENT CONTROL
AND WRITTEN STANDARD OPERATING PROCEDURES
                    F-l                                ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               F-2

-------
             Exhibit  F  -  Chain-of-Custody,  Document  Control  and
                    Written Standard Operating Procedures

                               Table of Contents

Section                                                                    Page

1.0   INTRODUCTION  	   5
      1.1   Purpose of Evidence Requirements   	  5

2.0   STANDARD OPERATING PROCEDURES 	   6
      2.1   Sample Receiving   	  6
      2.2   Sample Identification	7
      2.3   Sample Security  	   7
      2.4   Sample Storage   	  7
      2.5   Sample Tracking and Document Control   	  8
      2.6   Computer-Resident Sample Data Contro 1 	  9
      2.7   Complete SDG File  (CSF) Organization and Assembly  	   9

3.0   WRITTEN STANDARD OPERATING PROCEDURES 	 11
      3.1   Sample Receiving   	 11
      3.2   Sample Identification	12
      3.3   Sample Security	.'	13
      3.4   Sample Storage   	13
      3.5   Sample Tracking and Document Control   	  13
      3.6   Computer-Resident Sample Data Contro 1 	  14
      3.7   CSF Organization and Assembly	15
                                       F-3                               ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               F-4

-------
                                                         Exhibit F — Section 1
                                                                   Introduction
1.0   INTRODUCTION
      A sample is physical evidence collected from a facility or from the
      environment.  Controlling evidence is an essential part of the hazardous
      waste investigation effort.  To ensure that U.S. Environmental
      Protection Agency's (USEPA's) sample data and records supporting sample-
      related activities are admissible and have weight as evidence in future
      litigation. Contractors are required to maintain USEPA samples under
      chain-of-custody and to account for all samples and supporting records
      of sample handling, preparation, and analysis.  Contractors shall
      maintain sample identity, sample custody, and all sample-related records
      according to the requirements in this exhibit.

1.1   Purpose of Evidence Requirements

      The purpose of the evidence requirements include:

      =" Ensuring traceability of  samples while in possession of the
         Contractor;

      '•** Ensuring custody of samples while in possession of the Contractor;

      :xl Ensuring the integrity of sample identity while in possession of the
         Contractor;

      '-*1 Ensuring sample-related activities are recorded on documents or in
         other formats for USEPA sample receipt, storage, preparation,
         analysis, and disposal;

      ™ Ensuring all laboratory records for each specified Sample Delivery
         Group will be accounted for when the project  is completed; and

      !in Ensuring that all laboratory records directly related to USEPA
         samples  are assembled and delivered to USEPA  or, prior to delivery,
         are  available upon USEPA's request.
                                       F-5                               ILM05.2

-------
Exhibit F — Section 2
Standard Operating Procedures

2.0   STANDARD OPERATING PROCEDURES

      The Contractor shall implement the following Standard Operating
      Procedures  (SOPs) for sample receiving, sample identification, sample
      security, sample storage, sample tracking and document control,
      computer-resident sample data control, and Complete Sample Delivery
      Group  (SDG) File  (CSF) organization and assembly to ensure
      accountability of USEPA sample chain-of-custody as well as control of
      all USEPA sample-related records.

2.1   Sample Receiving

2.1.1    The Contractor shall designate a sample custodian responsible  for
         receiving USEPA samples.

2.1.2    The Contractor shall designate a representative to receive USEPA
         samples  in the event that the sample custodian is not  available.

2.1.3    Upon receipt, the condition of shipping containers and sample
         containers shall be inspected and  recorded on Form DC-1 by the sample
         custodian or a designated representative.

2.1.4    Upon receipt, the condition of the custody seals  (intact/broken)
         shall be inspected and recorded on Form DC-1 by the sample custodian
         or a designated representative.

2.1.5    The sample custodian or a designated representative shall verify and
         record on Form DC-1 the agreement  or disagreement of information
         recorded on all documents received with samples and information
         recorded on sample containers.

2.1.6  •  The sample custodian or a designated representative shall verify and
         record the following information on Form DC-1 as samples are received
         and inspected:

         **> Presence or absence and condition of custody seals  on shipping
      „•-,     and/or sample containers;

         :x: Custody seal numbers when present;

         '•x* Presence or absence of Traffic  Reports/Chain of Custody Records  or
            Packing Lists;

         :i: Presence or absence of airbills or airbill stickers;

         :x: Airbill or airbill sticker numbers;

         1X3 Presence or absence of sample tags;

         :*: Sample tags listed/not listed on Traffic Reports/Chain of Custody
            Records;

         00 Condition of the sample bottles;

         ^ Presence or absence of cooler temperature indicator bottle;

         :*: Cooler temperature;

         ^ Date  of receipt;

         :l: Time  of receipt;

         '*' EPA sample numbers;

ILM05.2                               F-6

-------
                                                         Exhibit F — Section 2
                                          Standard Operating Procedures (Con't)

         '•*=  pH of all aqueous samples;

         '•x*  Sample tag numbers;

         :*:  Assigned laboratory numbers;
                                >
         "=  Remarks regarding condition of sample shipment, etc.;

         <**  Samples delivered by hand; and

         :i:  Problems and discrepancies.

2.1.7    The sample custodian or a designated representative shall sign, date,
         and record the time on all accompanying forms, when applicable, at
         the time of sample receipt (e.g.. Traffic Reports/Chain of Custody
         Records or packing lists, and airbills).

         NOTE: Initials are not acceptable.

2.1.8    The Contractor shall contact the Sample Management Office  (SMO) to
         resolve problems and discrepancies including, but not limited  to:
         absent documents; conflicting information; absent or broken custody
         seals; insufficient sample volume; unsatisfactory sample condition
         (e.g., leaking sample container); and  samples not preserved to the
         proper pH.

2.1.9    The Contractor shall record the resolution of all problems and
         discrepancies communicated through SMO.

2.2   Sample Identification

2.2.1    The Contractor shall maintain the identity of USEPA samples and
         prepared samples  (including extracted  samples, digested samples, and
         distilled samples) throughout the laboratory.

2.2.2    Each sample and sample preparation container shall be labeled  with
         the EPA sample number or a unique laboratory sample identification
         number.

2.3   Sample Security

2.3.1    The Contractor shall demonstrate that  USEPA sample custody is
         maintained  from receiving through retention or disposal.  A sample is
         in custody  if:

         :x: It is in your possession; or

         =" It is in your view after being in your possession; or

         :i: It is locked in a  secure  area after being  in  your possession; or

         ^ It is in a designated secure  area.   (Secure areas  shall be
            accessible only to authorized personnel).

2.3.2    The Contractor shall demonstrate security of designated secure areas.

2.4   Sample Storage

      The Contractor shall designate  storage areas for USEPA samples and
      prepared samples.
                                       F-7                                ILM05.2

-------
Exhibit F — Section 2
Standard Operating Procedures (Con't)

2.5   Sample Tracking and Document Control

2.5.1    The Contractor shall record all activities performed on USEPA
         samples.

2.5.2    Titles which identify the activities recorded shall be printed on
         each page of all laboratory documents.   (Activities include, but are
         not limited to: sample receipt; sample storage; sample preparation,
         and sample analysis.)  When a document is a record of analysis, the
         instrument type and parameter group  [e.g., ICP-AES  (metals)] shall be
         included in the title.

2.5.3    When columns are used to organize information recorded on laboratory
         documents, the information recorded in the columns shall be
         identified in a column heading.

2.5.4    Reviewers' signatures shall be identified on laboratory documents
         when reviews are conducted.

         NOTE: Individuals recording review comments on computer-generated raw
         data are not required to be identified unless the written comments
         address data validity.

2.5.5    The laboratory name shall be identified on preprinted laboratory
         documents.

2.5.6    Each laboratory document entry shall be dated with the month/day/year
         (e.g., 01/01/1999) and signed by the individual(s) responsible for
         performing the recorded activity at the time the activity is
         recorded.

2.5.7    Notations on laboratory documents shall be recorded in ink.

2.5.8    Corrections to laboratory data reporting forms and raw data shall be
         made by drawing single lines through the errors and entering the
         correct information.  Information shall not be obliterated or
         rendered unreadable.  Corrections and additions to information shall
         be signed  (or initialed) and dated.

2.5.9    Unused portions of laboratory documents shall be lined-out.

2.5.10   Pages in bound and unbound logbooks shall be sequentially numbered.

2.5.11   Instrument-specific run logs shall be maintained to enable the
         reconstruction of run sequences.

2.5.12   Logbook entries shall be in chronological order.

2.5.13   Logbook entries shall include only one SDG per page, except in the
         events where SDGs "share" Quality Control (QC) samples  (e.g.,
         instrument run logs and extraction logs).

2.5.14   Each page in bound and unbound logbooks shall be dated
         (month/day/year) and signed  (no initials) at the bottom by the
         individual recording the activity  (if a single entry is made on a
         page) or by the last individual recording information on the page  (if
         multiple entries are on the same page).

2.5.15   Information inserted into laboratory documents shall be affixed
         permanently in place.  The individual responsible for inserting
         information shall sign and date across the insert and logbook page at
         the time information is inserted.
ILM05.2                               F-E

-------
                                                         Exhibit F — Section 2
                                          Standard Operating Procedures (Con't)

2.5.16   The Contractor shall document disposal or retention of USEPA samples,
         remaining portions of samples, and prepared samples.

2.6   Computer-Resident Sample Data Control

2.6.1    Contractor personnel responsible for original data entry shall be
         identified at the time of data input.

2.6.2    The Contractor shall make changes to electronic data in a manner
         which ensures that the original data entry is preserved, the editor
         is identified, and the revision date is recorded.

2.6.3    The Contractor shall routinely verify the accuracy of manually
         entered data, electronically entered data, and data acquired from
         instruments.

2.6.4    The Contractor shall routinely verify documents produced by the
         electronic data collection system to ensure accuracy of the
         information reported.

2.6.5    The Contractor shall ensure that the electronic data collection
         system is secure.

2.6.5.1     The electronic data collection system shall be maintained in a
            secure location.

2.6.5.2     Access to the electronic data collection system  functions shall be
            limited to authorized personnel  through utilization  of software
            security techniques  (e.g., log-ons or restricted passwords).

2.6.5.3     Electronic data collection systems shall be protected from the
            introduction of external programs  or  software  (e.g., viruses).

2.6.6    The Contractor shall designate archive storage areas for electronic
         data and the  software required to access the data.

2.6.7    The Contractor shall designate an individual responsible for
         maintaining archives of electronic  data  including  the software.

2.6.8    The Contractor shall maintain the archives of electronic data and
         necessary software in a secure location.   (Secure  areas shall be
         accessible only to authorized personnel.)

2.7   Complete SDG File  (CSF) Organization and Assembly

2.7.1    The Contractor shall designate a document control  officer responsible
         for the organization and assembly of  the CSF.

2.7.2    The Contractor shall designate a representative responsible for the
         organization  and  assembly  of  the CSF  in  the event  that  the document
         control officer is not available.

2.7.3    The Contractor shall maintain documents  relating to the CSF in a
         secure location.

2.7.4    All original  laboratory forms and copies of SDG-related logbook pages
         shall be  included in the CSF.

2.7.5    Copies of laboratory documents in the CSF shall be  photocopied in a
         manner to provide complete and  legible replicates.
                                       F-9                               ILM05.2

-------
Exhibit F — Section 2
Standard Operating Procedures (Con't)
2.7.6
2.7.7
2.7.8
2.7.!
Documents relevant to each SDG including, but not limited to, the
following shall be included in the CSF:
            logbook pages;
            bench sheets;
            screening records;
            preparation records;
            repreparation records;
            analytical records;
            re-analysis records;
            records of failed or
            attempted analysis;
                                          custody  records;
                                          sample tracking  records;
                                          raw  data summaries;
                                          computer printouts;
                                          correspondence;
                                          FAX  originals;
                                          library  search results; and
                                          other.
The document control officer or a designated' representative shall
ensure that sample tags are encased in clear plastic bags before
placing them in the CSF.

CSF documents shall be organized and assembled on an SDG-specific
basis.

Original documents which include information relating to more than
one SDG (e.g., Traffic Reports/Chain of Custody Records,- calibration
logs) shall be filed in the CSF of the lowest SDG number, and copies
of these originals shall be placed in the other CSF(s).  The document
control officer or a designated representative shall record the
following statement on the copies in  (indelible)  dark ink:

                             COPY
       ORIGINAL DOCUMENTS ARE INCLUDED IN CSF 	
                                                             Signature
                                                              Date

2.7.10   All CSFs shall be submitted with a completed Form DC-2.  All
         resubmitted CSFs shall be submitted with a new or revised  Form DC-2.

2.7.11   Each item in the CSF and resubmitted CSFs shall be inventoried and
         assembled in the order specified on Form DC-2.  Each page  of the CSF
         shall be stamped with a sequential number.  Page number ranges shall
         be recorded in the columns provided on  Form DC-2.  Intentional gaps
         in the page numbering sequence  shall be recorded in  the "Comments"
         section on Form DC-2.  When inserting new or inadvertently omitted
         documents, the Contractor shall identify them with unique  accountable
         numbers.  The unique accountable numbers and  the locations of the
         documents shall be recorded in  the "Other Records" section on Form
         DC-2.

2.7.12   Before shipping each CSF, the document  control officer or  a
         designated representative shall verify  the agreement of information
         recorded on all documentation and ensure that the information is
         consistent and the CSF is complete.

2.7.13   The document control officer or a designated representative shall
         document the shipment of deliverable packages including what was
         sent, to whom, the date, and the carrier used.

2.7.14   Shipments of deliverable packages, including resubmittals, shall be
         sealed with custody  seals by the document control officer  or a
ILM05.2
                                      F-10

-------
                                                    Exhibit F — Sections 2 & 3
                                          Written Standard Operating Procedures

         designated representative in a manner such that opening the packages
         would break the seals.

2.7.15   Custody seals shall be signed and dated by the document control
         officer or a designated representative when sealing deliverable
         packages.

3.0   WRITTEN STANDARD OPERATING PROCEDURES

      The Contractor shall develop and implement the following written
      Standard Operating Procedures  (SOPs) for sample receiving, sample
      identification, sample security, sample storage, sample tracking and
      document control, computer-resident sample data control, and Complete
      Sample Delivery Group  (SDG) File  (CSF) organization and assembly to
      ensure accountability for USEPA sample chain-of-custody and control of
      all USEPA sample-related records.

3.1   Sample Receiving

3.1.1    The Contractor shall have written SOPs for sample receiving which
         accurately reflect the procedures used by the laboratory.

3.1.2    The written SOPs for sample receiving shall ensure that the
         procedures listed below are in use at the laboratory.

3.1.2.1     The condition of shipping containers and sample containers are
            inspected and recorded on Form  DC-1 upon receipt by the  sample
            custodian or a designated representative.

3.1.2.2     The condition of custody seals are inspected and recorded on Form
            DC-1  upon receipt  by the sample custodian  or  a designated
            representative.

3.1.2.3     The presence or absence of  the  following documents/items
            accompanying the sample  shipment  is verified  and recorded on Form
            DC-1  by  the sample  custodian  or a designated  representative:

            00  Custody  seals;

            !*:  Traffic  Reports/Chain of Custody  Records or Packing Lists;

            "=  Airbills or  airbill  stickers;

            :*:  Sample  tags;  and

            :x:  Cooler  temperature  indicator bottle.

3.1.2.4     The agreement or disagreement of  information  recorded  on shipping
            documents with  information  recorded  on  sample  containers is
            verified and recorded on Form DC-1 by the  sample custodian  or  a
            designated  representative.

3.1.2.5     The following information is  recorded on Form DC-1 by  the sample
            custodian  or a  designated  representative  as samples are received
            and inspected:

            :x:  Custody seal numbers,  when present;

            ^  Airbill  or  airbill  sticker numbers;

            :*:  Sample tag  numbers  listed/not  listed on Traffic  Reports/Chain
                of Custody  Records;


                                      F-ll                               ILM05.2

-------
Exhibit F — Section 3
Written Standard Operating Procedures  (Con't)

            30 Condition  of  sample  bottles;

            1X2 Cooler  temperature;

            :x: Date of receipt;

            =** Time of receipt;

            <*> EPA sample numbers;

            !M! pH of all  aqueous  samples;

            = Sample  tag numbers;

            =*= Assigned laboratory  numbers;

            !x: Remarks regarding  condition of  sample shipment,  etc.;

            00 Samples delivered  by hand; and

            '•** Problems and  discrepancies.

3.1.2.6     All accompanying forms  are signed, dated, and the time is
            recorded,  when applicable, at the  time  of sample receipt  (e.g.,
            Traffic Reports/Chain of Custody Records or packing lists,  and
            airbills)  by  the sample custodian  or  a  designated representative.

3.1.2.7     The Sample  Management Office  (SMO) is contacted to  resolve
            problems and  discrepancies including, but not limited to: absent
            documents;  conflicting  information; absent  or broken custody
            seals; insufficient sample volume; unsatisfactory sample  condition
             (e.g., leaking sample container);  and samples not preserved to the
            proper pH.

3.1.2.8     The resolution of all problems and discrepancies communicated
            through SMO is recorded.

3.2   Sample Identification

3.2.1    The Contractor shall have  written SOPs for sample identification
         which accurately reflect the procedures  used by the laboratory.

3.2.2    The written SOPs for sample identification shall ensure that the
         procedures listed below  are in use  at the  laboratory.

3.2.2.1     The identity  of USEPA samples and  prepared  samples  is maintained
            throughout the laboratory when:

            <** The Contractor assigns unique laboratory sample  identification
               numbers, the  written SOPs  shall include  a description  of the
               procedure  used to  assign these  numbers;

            ™ The Contractor uses  prefixes  or suffixes in addition to
               laboratory sample  identification numbers, the written  SOPs
               shall include their  definitions; and

            '•*'• The Contractor uses  methods to  uniquely  identify
               fractions/parameter  groups and  matrix type, the  written  SOPs
               shall include a description of  these methods.

3.2.2.2     Each sample and sample  preparation container is labeled with the
            SMO number or a unique  laboratory  sample identification number.


ILM05.2                               F-12

-------
                                                         Exhibit F — Section  3
                                  Written Standard Operating Procedures  (Con't)
3.3   Sample Security
3.3.1    The Contractor shall have written SOPs for sample security which
         accurately reflect the procedures used by the laboratory.

3.3.2    The written SOPs for sample security shall include the items listed
         below.

3.3.2.1     Procedures which ensure the following:

            :i: Sample custody is maintained; and

            =*• The security of designated secure  areas is maintained.

3.3.2.2     A list of authorized personnel who have access to locked storage
            areas.

3.4   Sample Storage

3.4.1    The Contractor shall have written SOPs for sample storage which
         accurately reflect the procedures used by the laboratory.

3.4.2    The written SOPs for sample storage shall describe locations,
         contents, and identities of all storage  areas for USEPA  samples  and
         prepared samples in the laboratory.

3.5   Sample Tracking and Document Control

3.5.1    The Contractor shall have written SOPs for sample tracking and
         document control which accurately reflect the procedures used  by the
         laboratory.

3.5.2    The written SOPs for sample tracking and document control shall
         include the items listed below.

3.5.2.1     Examples of all laboratory  documents  used during  sample receiving,
            sample  storage,  sample  transfer,  sample analyses, CSF organization
            and assembly, and sample retention or disposal.

3.5.2.2     Procedures which ensure the following:

            =*= All  activities performed on USEPA  samples are  recorded;

            :x: Titles  which  identify the activities recorded  are  printed on
               each page  of  all  laboratory documents;

            '•*• Information recorded in  columns  is identified  with column
               headings;

            ;x: Reviewers'  signatures are  identified on  laboratory documents;

            ^ The  laboratory  name  is  included on preprinted  laboratory
               documents;

            00 Laboratory document  entries  are  signed and dated with the
               month/day/year  (e.g.,  01/01/1999);

            '•*• Entries on all  laboratory  documents  are  recorded in  ink;

            '•*> Corrections and additions  to  laboratory  documents  are made by
               drawing single  lines through  the  errors,  entering  the correct
               information,  and initialing  and dating the new information;


                                      F-13                               ILM05.2

-------
Exhibit F — Section 3
Written Standard Operating Procedures  (Con't)

            a! Unused portions  of laboratory documents  are lined-out;

            =" Pages in  bound and unbound logbooks  are  sequentially numbered;

            !X! Instrument-specific run logs  are maintained to enable the
               reconstruction of run sequences;

            '•*•• Logbook entries  are recorded  in chronological order;

            :*: Entries are  recorded for only one SDG on a page,  except in the
               event where  SDGs "share" Quality Control (QC)  samples (e.g.,
               instrument run logs and extraction logs);

            =•* Each page in bound and unbound logbooks  shall be  dated
                (month/day/year)  and signed (no initials)  at the  bottom by the
               individual recording the activity (if a  single entry is made  on
               a page) or by the last individual recording information on the
               page  (if  multiple entries are on the same  page);

            =•= Information  inserted in laboratory documents is affixed
               permanently,  signed,  and dated across the  insert;  and

            :*: The retention or disposal of  USEPA samples,  remaining portions
               of samples,  and  prepared samples is  documented.

3.6   Computer-Resident  Sample  Data  Control

3.6.1    The Contractor  shall have written SOPs for computer-resident  sample
         data control which  accurately reflect the  procedures used by  the
         laboratory.

3.6.2    The written SOPs for computer-resident sample  data control shall
         include the items  listed below.
                                       i
3.6.2.1     Procedures which ensure  the following:

            '•*= Contractor personnel  responsible for original data entry are
               identified;

            :x: Changes to electronic data are made  such that the original data
               entry is  preserved, the editor is identified,  and the revision
               date is recorded;

            :M: The accuracy of  manually entered data, electronically entered
               data, and data acquired from  instruments is verified;

            ™ Report documents produced by  the electronic data  collection
               system are routinely verified to ensure  the accuracy of the
               information  reported;

            :I: Electronic data  collection system security is maintained;

            a3 Archives  of  electronic data and accompanying software are
               maintained in a  secure location; and

            :*! Off-site  backup  and storage of electronic  data is maintained.

3.6.2.2     Descriptions of  archive  storage  areas for the electronic data and
            the software required to access  data archives.

3.6.2.3     A list of authorized personnel who have access  to electronic data
            collection system functions and  to archived data.


ILM05.2                               F-14

-------
                                                         Exhibit F — Section 3
                                  Written Standard Operating Procedures  (Con't)

3.7   CSF Organization and Assembly

3.7.1    The Contractor shall have written SOPs for CSF organization and
         assembly which accurately reflect the procedures used by the
         laboratory.

3.7.2    The written SOPs for CSF organization and assembly shall ensure that
         the procedures listed below are in use at the laboratory.

         ••*•• Documents relating to the CSF are maintained in a secure location.

         ••*•• All original laboratory forms and copies of SDG-related logbook
            pages are included in the CSF.

         =*= Laboratory documents are photocopied  in a manner to provide
            complete and legible replicates.

         00 All documents relevant to each  SDG are included in the CSF.

         '•x* Sample  tags are  encased in  clear plastic bags by the document
            control officer  or a designated representative before placing  them
            in the  CSF.

         ™ The CSF is organized and assembled on an SDG-specific basis.

         ^ Original documents which contain information relating to more  than
            one SDG are filed in the CSF of the lowest SDG  and copies are
            referenced to originals in  the  event  that an original document
            contains information  relating to more than one  SDG.

         =ť Each CSF is submitted with  a completed Form DC-2, and resubmitted
            CSFs are submitted with a new or revised Form DC-2.

         :l! Each page of the CSF  is stamped with  a sequential number and the
            page number ranges are recorded in the columns  provided on  Form
            DC-2.   Intentional gaps in  the  page numbering sequence are
            recorded in the  "Comments"  section of Form DC-2.  Inserted
            documents are recorded in the "Other  Records" section of Form  DC-
            2.

         =*•• Consistency and  completeness of the CSF are verified by the
            document control officer or a designated representative.

         :*: Shipments of deliverable packages are documented by the document
            control officer  or a  designated representative.

         "•• Deliverable packages  are shipped by the document control, officer
            or  a designated  representative  using  custody seals in a manner
            such that opening the packages  would  break the  seals.

         :i: Custody seals are signed and dated by the document control  officer
            or  a designated  representative  before placing them on deliverable
            packages.
                                      F-15                               ILM05.2

-------
THIS PAGE INTENTIONALLY LEFT BLANK

-------
    EXHIBIT G




GLOSSARY OF TERMS
       6-1                               ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               G-2

-------
                                                 Exhibit G -- Glossary of Terms

ABSORBANCE - A measure of the decrease in incident light passing through a
sample into a detector.  It is defined mathematically as:


            Absorbance


                                  A = -log 	
                                           lo


            WHERE,

                   I  = Radiation  intensity of a sample.
                   IB  = Radiation intensity of  a blank.

ALIQUOT - A measured  portion of a field sample, standard, or solution taken
for sample preparation and/or analysis.

ANALYSIS DATE/TIME -  The date and military time  (24-hour clock) of the
introduction  of  the  sample,  standard,  or  blank into  the  analysis  system.

ANALYTE - The element or ion an analysis  seeks to determine; the element of
interest.

ANALYTICAL SAMPLE - Any solution or media introduced into an instrument on
which an analysis is  performed, excluding instrument calibration, initial
calibration verification  (ICV), initial calibration  blank  (ICB), continuing
calibration verification  (CCV), continuing calibration blank  (CCB),  and tunes.
Note the following are all defined as  analytical samples: undiluted  and
diluted samples  (USEPA and non-USEPA), matrix  spike  samples, duplicate
samples, serial  dilution samples, analytical spike samples, post-digestion
spike samples. Interference  Check Samples  (ICSs), Contract Required
Quantitation  Limit  (CRQL) Check Standards  (CRIs), Laboratory Control Samples
 (LCSs), Performance  Evaluation  (PE) samples, Preparation Blanks  (PBs), and
Linear Range  Samples  (LRSs).

ANALYTICAL SEQUENCE  - The actual instrumental  analysis of the  samples from the
time of instrument calibration through the analysis  of the final CCV or CCB.
All sample analyses  during the analytical  sequence are subject  to the QC
protocols set forth  in Exhibits D and  E of this contract unless otherwise
specified in  the individual  methods.

ANALYTICAL SPIKE - A spike that is fortified just prior  to analysis  by adding
a  known quantity of  the analyte to an  aliquot  of the prepared  sample.

ASTM - American  Society for  Testing and Materials.   A developer and  provider
of voluntary  consensus standards.

AUTOZERO - Zeroing the instrument at the  proper wavelength.   It is equivalent
to running a  standard blank  with the absorbance  set  at zero.

BACKGROUND CORRECTION - A technique to compensate for variable  background
contribution  to  the  instrument signal  in  the determination of  trace  elements.

BATCH - A group  of samples prepared at the same  time in  the same  location
using the same method.

BLANK - An analytical sample designed  to  assess  specific sources  of
contamination.   See  individual definitions for types of  blanks.

CALIBRATION  - The  establishment  of an  analytical curve based  on the
absorbance,  emission intensity,  or other  measured characteristic  of  known
                                       G-3                               ILM05.2

-------
Exhibit G — Glossary of Terms  (Con't)

standards.  The calibration standards must be prepared using the same type of
reagents or concentration of acids as used in the sample preparation.

CALIBRATION BLANK - A blank solution containing all of the reagents and in the
same concentration as those used in the analytical sample preparation.  This
blank is not subjected to the preparation method.

CALIBRATION STANDARDS - A series of known standard solutions used by the
analyst for calibration of the instrument (i.e., preparation of the analytical
curve).  The solutions may or may not be subjected to the preparation method
but contain the same matrix (i.e., the same amount of reagents and/or
preservatives) as the sample preparations to be analyzed.

CASE - A finite, usually predetermined number of samples collected over a
given time period from a particular site.  Case numbers are assigned by the
Sample Management Office  (SMO).  A Case consists of one or more Sample
Delivery Groups (SDGs).

CONCENTRATION LEVEL  (low or medium) - For inorganics analysis, low or medium
level is defined by the appropriate designation by the sampler on the Traffic
Report/Chain of Custody Record.

CONTAMINATION - A component of a sample or an extract that is not
representative of the environmental source of the sample.  Contamination may
stem from other samples, sampling equipment, while in transit, from laboratory
reagents laboratory environment, or analytical instruments.

CONTINUING CALIBRATION VERIFICATION  (CCV) - A single parameter or multi-
parameter standard solution prepared by the analyst and used to verify the
stability of the instrument calibration with time, and the instrument
performance during the analysis of samples.  The CCV can be one of the
calibration standards.  However, all parameters being measured by the
particular system must be represented in this standard and the standard must
have the same matrix  (i.e., the same amount of reagents and/or preservatives)
as the samples.  The CCV should have a concentration in the middle of the
calibration range and shall be run every 10 analytical samples or every 2
hours, whichever is more frequent.

CONTRACT COMPLIANCE SCREENING (CCS) - A screening of electronic and hardcopy
data deliverables for completeness and compliance with the contract.  This
screening is done under USEPA direction by the SMO Contractor.

CONTRACT LABORATORY PROGRAM (CLP) - Supports the USEPA's Superfund effort by
providing a range of state-of-the-art chemical analytical services of known
quality.  This program is directed by the Analytical Operations/Data Quality
Center  (AOC) of the Office of Emergency and Remedial Response  (OERR) of USEPA.

CONTRACT REQUIRED QUANTITATION LIMIT  (CRQL) - Minimum level of quantitation
acceptable under the contract Statement of Work  (SOW).

CONTRACT REQUIRED QUANTITATION LIMIT  (CRQL) CHECK STANDARD  (CRI) - A single
parameter or multi-parameter standard solution prepared at the CRQL and used
to verify the instrument calibration at low levels.

CONTROL LIMITS - A range within which specified measurement results must fall
to be compliant.  Control limits may be mandatory, requiring corrective action
if exceeded, or advisory, requiring that noncompliant data be flagged.

CYANIDE (Total)  - Cyanide ion and complex cyanides converted to hydrocyanic
acid  (HCN) by reaction in a reflux system of a mineral acid in the presence of
magnesium ion.
ILM05.2                                G-4

-------
                                         Exhibit G — Glossary of Terms  (Con't)

DATE - MM/DD/YYYY - Where MM = 01 for January, 02 for February, ... 12 for
December;  DD = 01 to 31; YYYY = 1998, 1999, 2000, 2001, etc.

DAY - Unless otherwise specified, day shall mean calendar day.

DIGESTION LOG - An official record of the sample preparation  (digestion).

DIRECT ANALYSIS - Analysis of a sample, standard, or blank that has not been
taken through a preparation procedure  (digestion or distillation).

DISSOLVED METALS - Analyte elements in a water/aqueous sample which will pass
through a 0.45 micrometer  (um) filter.

DRY WEIGHT - The weight of a sample based on percent solids.  The weight after
drying in an oven.

DUPLICATE - A second aliquot of a sample that is treated the  same as the
original sample in order to determine the precision of the method.

FIELD BLANK - This is any sample that is submitted from the field and is
identified as a blank.  This includes trip blanks, rinsates,  equipment blanks,
etc.

FIELD QC - Any Quality Control sample submitted  from the field to the
laboratory.  Examples include, but are not limited to: field  blanks, field
duplicates, and field spikes.

FIELD SAMPLE - A portion of material received to be analyzed  that is contained
in single or multiple containers and identified by a unique EPA sample number.

GRAPHITE FURNACE ATOMIC ABSORPTION  (GFAA) - A technique for the determination
of analytes in which a  sample aliquot is injected into a hollow graphite tube,
which is then heated to atomize  the analyte.  The vapor absorbs light at
wavelengths characteristic of the element(s) atoms present.

HOLDING TIME - The elapsed time  expressed in days from the date of  receipt of
the sample by the Contractor until the date of its analysis.

          Holding time =  (sample analysis date - sample receipt date)

INDEPENDENT STANDARD -  A Contractor-prepared  standard  solution that is
composed of analytes from  a different  source  than those used  in the standards
for the calibration.

INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROSCOPY  (ICP-AES)  - A
technique  for the simultaneous or sequential multi-element determination of
elements in solution.   The basis of  the method is the measurement  of atomic
emission by an optical  spectroscopic technique.  Characteristic atomic  line
emission spectra  are produced by excitation of the sample in  a radio frequency
inductively coupled plasma.

INDUCTIVELY COUPLED PLASMA-MASS  SPECTROSCOPY  (ICP-MS)  - A technique for  the
multi-element determination of elements in solution.   The basis of  the
technique  is the  detection of atomic ions produced by  an ICP  and  sorted  by
mass/charge ratio.

IN-HOUSE - At the Contractor's facility.

INITIAL CALIBRATION - Analysis of analytical  standards  for a  series of
different  specified concentrations;  used to define the  quantitative response,
linearity, and dynamic  range  of  the  instrument to target analytes.
                                       G-5                               ILM05.2

-------
Exhibit G — Glossary of Terms  (Con't)

INITIAL CALIBRATION VERIFICATION  (ICV) - Solution(s) prepared from stock
standard solutions, metals or salts obtained from a source separate from that
utilized to prepare the calibration standards.  The ICV is used to verify the
concentration of the calibration standards and the adequacy of the instrument
calibration.  The ICV should be traceable to NIST or other certified standard
sources when USEPA ICV solutions are not available.

INJECTION - Introduction of the analytical sample into the instrument
excitation system for the purpose of measuring absorbance, emission or
concentration of an analyte.  May also be referred to as exposure  .

INSUFFICIENT QUANTITY - When there is not enough volume (water/aqueous sample)
or weight (soil/sediment) to perform any of the required operations: sample
analysis, percent solids, etc.  Exhibit D provides guidance for addressing
this problem.

INTERFERENCE CHECK SAMPLE - A solution containing both interfering and analyte
elements of known concentration that can be used to verify background and
interelement correction factors.

INTERFERENTS - Substances which affect the analysis for the element of
interest.

INTERNAL STANDARD - A non-target element added to a sample at a known
concentration after preparation but prior to analysis.  Instrument responses
to internal standards are monitored as a means of assessing overall instrument
performance.

LABORATORY - Synonymous with Contractor as used herein.

LABORATORY CONTROL SAMPLE (LCS) - A control sample of known composition.
Laboratory control samples are analyzed using the same sample preparation,
reagents, and analytical methods employed for the USEPA samples received.

LABORATORY RECEIPT DATE - The date on which a sample is received at the
Contractor's facility, as recorded on the shipper's delivery receipt and
Sample Traffic Report/Chain of Custody Record.  Also referred to as VTSR
(Validated Time of Sample Receipt).

LINEAR RANGE, LINEAR DYNAMIC RANGE - The concentration range over which the
instrument response remains linear.

MATRIX - The predominant material of which the sample to be analyzed is
composed.  For the purpose of this SOW, a sample matrix is either water/
aqueous or soil/sediment.  Matrix is not synonymous with phase  (liquid or
solid).

MATRIX EFFECT - In general, the effect of particular matrix constituents.

MATRIX SPIKE - Aliquot of a sample  (water/aqueous or soil) fortified (spiked)
with known quantities of specific compounds and subjected to the entire
analytical procedure in order to indicate the appropriateness of the method
for the matrix by measuring recovery.

METHOD DETECTION LIMIT  (MDL) - The concentration of a target parameter that,
when a sample is processed through the complete method, produces a signal with
99 percent probability that it is different from the blank.  For 7 replicates
of the sample, the mean value must be 3.14s above the blank, where "s" is the
standard deviation of the 7 replicates.

NARRATIVE (SDG Narrative) - Portion of the data package which includes
laboratory, contract, Case, sample number identification, and descriptive
documentation of any problems encountered in processing the samples, along


ILM05.2                                G-6

-------
                                         Exhibit G — Glossary of Terms  (Con't)

with corrective action taken and problem resolution.  Complete SDG Narrative
specifications are included in Exhibit B.

PERCENT DIFFERENCE (%D) - As used in this SOW and elsewhere to compare two
values.  The difference between the two values divided by one of the values.

PERCENT SOLIDS (%S) - The proportion of solid in a soil sample determined by
drying an aliquot of the sample.

PERFORMANCE EVALUATION (PE) SAMPLE - A sample of known composition provided by
USEPA for Contractor analysis.  Used by USEPA to evaluate Contractor
performance.

PREPARATION BLANK - An analytical control that contains reagent water and
reagents, which is carried through the entire preparation and analytical
procedure.

PREPARATION LOG - An official record of the sample preparation  (digestion or
distillation).

QUALITY ASSURANCE TECHNICAL SUPPORT  (QATS) LABORATORY - A Contractor-operated
facility operated under the QATS contract, awarded and administered by USEPA.

REAGENT WATER - The purity of this water must be equivalent to ASTM Type II
reagent water of Specification D1193-77, "Standard Specification for Reagent
Water".

RELATIVE PERCENT DIFFERENCE  (RPD) - As used in this SOW and elsewhere to
compare two values, the relative percent difference is based on the mean of
the two values, and is reported as an absolute value, i.e., always expressed
as a positive number or zero.

REPRESENTATIVE - Alternate or designee who has the knowledge and authority to
perform a specific task.

ROUNDING RULES - If the figure is greater than or equal to 5, round up,
otherwise round down.  As  an example, 11.443 is rounded down to 11.44 and
11.455 is rounded up to 11.46.  If a series of multiple operations is to be
performed  (add, subtract,  divide, multiply), all figures are carried through
the calculations.  Then the  final answer is rounded to the proper number of
significant  figures.   See  forms instructions  (Exhibit B) for exceptions.

RUN - A continuous analytical sequence consisting of prepared samples and
all associated Quality Assurance  (QA) measurements as required by the contract
SOW.  A run begins with the  instrument calibration and is to be completed
within a 24-hour period.

SAMPLE - A portion of  material to be analyzed that is contained in single or
multiple containers and identified by a unique sample number.

SAMPLE DELIVERY GROUP  (SDG)  - A unit within a sample Case that  is used to
identify a group of samples  for delivery.  An SDG is defined by the following,
whichever is  most  frequent:

"=     Each  Case of field  samples  received, or

™     Each  20 field  samples  [excluding  Performance Evaluation  (PE) samples]
       within a Case,  or

"=     Each  7 calendar day period  (3  calendar  day period  for  7 day turnaround)
       during which field  samples  in  a Case are received  (said period beginning
       with  the receipt of the  first  sample in the SDG).
                                       G-7                               ILM05.2

-------
Exhibit G — Glossary of Terms  (Con't)

IX!     In addition, all samples and/or sample fractions assigned to an SDG must
      have been scheduled under the same contractual turnaround time.
      Preliminary Results have  no impact on defining the SDG.

Samples may be assigned to SDGs by matrix  (i.e., all soil samples in one SDG,
all water samples in another) at the discretion of the laboratory.

SAMPLE MANAGEMENT OFFICE (SMO) - A Contractor-operated facility operated under
the SMO contract, awarded and administered by USEPA.

SAMPLE NUMBER  (EPA SAMPLE NUMBER) - A unique identification number
designated by USEPA for each sample.  The EPA sample number appears on the
sample Traffic Report/Chain of Custody Record which documents information on
that sample.

SENSITIVITY - The slope of the analytical curve  (i.e., functional relationship
between instrument response and concentration).

SERIAL DILUTION - The dilution of a sample by a factor of five.  When
corrected by the dilution factor, the diluted sample must agree with the
original undiluted sample within specified limits.  Serial dilution may
reflect the influence of interferents.

SOIL - Synonymous with soil/sediment or sediment as used herein.

SOP - Standard Operating Procedure.

SOW - Statement of Work.

STANDARD ANALYSIS - An analytical determination made with known quantities of
target analytes.

STOCK SOLUTION - A standard solution which can be diluted to derive other
standards.

TARGET ANALYTE LIST (TAL) - A list of Inorganic Analytes  (metals and cyanide)
as designated in Exhibit C.

TIME - When required to record time on any deliverable item, time shall be
expressed as Military Time [i.e., a 24-hour clock  (0000-2359)].

TRAFFIC REPORT/CHAIN OF CUSTODY RECORD  (TR/COC) - An USEPA sample
identification form filled out by the sampler, which accompanies the sample
during shipment to the laboratory and is used for documenting sample identity,
sample chain-of-custody, and sample receipt by the laboratory.

TUNE - Analysis of a solution containing a range of isotope masses to
establish ICP-MS accuracy, resolution, and precision prior to calibration.

USEPA OERR AOC INORGANIC PROGRAM MANAGER  (AOC PM) - The USEPA, OERR AOC
Official who manages the CLP Inorganic Program.

USEPA REGIONAL CLP PROJECT OFFICER  (CLP PO) - The Regional USEPA official
responsible for monitoring laboratory performance and/or requesting analytical
data or services from a CLP laboratory.

VALIDATED TIME OF SAMPLE RECEIPT  (VTSR) - The date on which a sample is
received at the Contractor's facility, as recorded on the shipper's delivery
receipt and Sample Traffic Report/Chain of Custody Record.

WET WEI'GHT - The weight of a sample aliquot including moisture  (undried).
 ILM05.2                                G-8

-------
                                         Exhibit G — Glossary of Terms (Con't)

10% FREQUENCY - A frequency specification during an analytical sequence
allowing for no more than 10 analytical samples between required calibration
verification measurements,  as specified by the contract SOW.
                                       G-9                               ILM05.2

-------
THIS PAGE INTENTIONALLY LEFT BLANK

-------
         EXHIBIT H

DATA DICTIONARY AND FORMAT
 FOR DATA DELIVERABLES IN
  COMPUTER-READABLE  FORMAT
             H-l                               ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               H-2

-------
         Exhibit H - Data Dictionary and Format for Data Deliverables
                          in Computer-Readable Format

                               Table of Contents
Section                                                                     Page

1.0   USEPA AGENCY STANDARD IMPLEMENTATION   	   5
      1.1   Format Characteristics   	   5

2.0   RECORD TYPES   	   7
      2.1   Specifications   	   7

3.0   PRODUCTION RUNS	8
      3.1   Specifications   	   8
      3.2   Example	8

4.0   RECORD SEQUENCE	11
      4.1   Specifications   	11

5.0   FILE/RECORD INTEGRITY 	 13

6.0   DATES AND TIMES	13

7.0   MULTIPLE VOLUME DATA   	13

8.0   DELIVERABLE	14
      8.1   Requirements   	14

9.0   RECORD LISTING   	15
      9.1   Production Run  First Header Record  (Type  10)    	  15
      9.2   Production Run  Second Header Record  (Type 16)  	  16
      9.3   Mandatory Sample Header Data Record  (Type 20  )  	  17
      9.4   Sample Header Record  (Type 21)   	  22
      9.5   Associated Injection and Counter Record (Type  22)  	 24
      9.6   Results  Data Record  (Type 30)	25
      9.7   Instrumental Data  Readout  (Type 31)	28
      9.8   Auxiliary Data  Record  (Type 32)	29
      9.9   QC Limit Record (Type 34 )	30
      9.10  Correction Data Record  (Type 35)   	31
      9.11  Comment  Record  (Type 90)   	32
      9.12  Sample Associated  Data Record  (Type  92  )	33
                                       H-3                               ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                               H-4

-------
                                                         Exhibit H — Section 1
                                           USEPA Agency Standard Implementation

1.0   USEPA AGENCY STANDARD IMPLEMENTATION

1.1   Format Characteristics

      The following constitutes an implementation  of the USEPA Agency Standard
      for Electronic Data Transmission based upon  analytical results and
      ancillary information required by the contract.  All data generated by a
      single analysis are grouped together, and the groups are aggregated to
      produce files that report data from a Sample Delivery Group (SDG).
      Because this implementation is only a subset of the USEPA Agency
      Standard, some fields have been replaced by  delimiters as place holders
      for non-Contract Laboratory Program  (CLP) data elements.

1.1.1    This implementation includes detailed specifications for the required
         format of each record.  The position in the record where each field
         is to be contained relevant to other fields is specified, as well as
         the maximum length of the field.  Each field's required contents are
         specified as literal  (contained in quotes), which must appear exactly
         as shown  (without quotes), or as a variable for which format and/or
         descriptions are listed in the format/contents column.  Options and
         examples are listed for most fields.  For fields where more than
         three options are available, a list and description of options are
         supplied following the record descriptions.  Fields are separated
         from each other by the delimiter "|"  (ASCII 124).  Fields that do not
         contain data should be zero length or a blank field  (empty with no
         space or additional delimiters between the delimiters before and
         after the field) with the delimiter as a place holder.  For the
         purposes of Section 9 of this Exhibit, wherever "blank" is given as
         an option under the "Format/Contents" column, it refers to a blank
         field as explained above.

1.1.2    Numeric fields may contain numeric digits, a decimal place, and a
         leading minus sign.  A positive sign  is assumed if no negative sign
         is entered in a numeric field and must not be entered into any
         numeric field.  Values that exceed the maximum length allowed shall
         be reported to the maximum possible, maintaining the specified
         decimal place and maximum field length restrictions.

1.1.3    Requirements for significant figures and number of decimal places are
         specified in Exhibit B.  The numeric  field  lengths are specified such
         that  all possible numeric values  can be written to the file.  The
         size  of the numeric field indicates the maximum number of digits,
         including a decimal place and negative sign,  if appropriate, that can
         appear in the field at the same time.  Therefore, the number reported
         may need to be rounded  (using rounding rules  described in Exhibit B)
         to fit into the field.  The rounding must maintain the greatest
         significance possible providing the field length limitation.  In
         addition, the rounded number that appears on  the form, and therefore
         the field in the diskette file, must  be used  in any  calculation that
         may result in other numbers reported  on the same form or other forms
         in the SDG.  Field lengths should only be as  long as necessary to
         contain the data; packing with blanks  is not  allowed.

1.1.4    USEPA is currently developing a data  delivery strategy that may be
         used  as an alternative to the requirements  stated in Exhibit H.  This
         strategy's intent  is  to provide a neutral data delivery structure to
         the Contractor that will  further  facilitate the exchange of
         analytical information generated  under this analytical protocol.  The
         proposed  strategy  is  intended to  accommodate  laboratories that
         generate data transmission files  under multiple data formats.  Upon
         implementation of  this alternate  electronic data delivery strategy by
         the USEPA and prior to submission of  data in  alternate'format(s), the


                                       H-5                                ILM05.2

-------
Exhibit H — Section 1
USEPA Agency Standard Implementation  (Con't)

         Contractor must first demonstrate  its ability to provide electronic
         data as stated in this Exhibit H and obtain written permission from
         the USEPA for the submission of data in alternate format(s).  The
         Contractor will receive a written  response to its request within 90
         calendar days.  However, until the implementation of this alternate
         electronic data delivery strategy by the USEPA, all electronic data
         deliverables must be provided as specified in this Exhibit H.
ILM05.2                               H-6

-------
                                                         Exhibit H — Section 2
                                                                   Record Types

2.0   RECORD TYPES

2.1   Specifications

      The USEPA Agency Standard consists of variable length ASCII records.
      Maximum field length specifications match the reporting requirements in
      Exhibit B.  The last two bytes of each record must contain "carriage
      return" and "line feed", respectively.

2.1.1    This implementation consists of twelve record types that can be
         summarized in four groups, designated by the first record type in
         each group:
          Type
          Run Header


          Sample Header

          Results Record
          Comments Record
Type ID   Contents
  10      Information pertinent to a group of
          samples processed in a continuous
          sequence; usually several per SDG
  20      Sample identifying,  qualifying,  and
          linking information
  30      Analyte results and qualifications
  90      Free form comments
2.1.2    All record types given are mandatory.  Type 10, representing the
         analytical run, contains the instrument and run IDs which act as an
         identifying label for the run.  All 10, 20, 30, and 90  series records
         following that record pertain to the same analytical run.  Type 20,
         representing the sample, contains the EPA Sample ID which acts as an
         identifying label for the sample.  The QC code indicates whether the
         data  is  from an environmental sample, calibration, or QC sample.  All
         20, 30,  and 90 series records following that  record pertain to the
         same  sample.  Type  30, representing'an individual analyte, contains
         an  identifier to identify the analyte.  All 30 series records
         following that record pertain to the same analyte.
                                       H-7
                                                                         ILM05.2

-------
Exhibit H — Section 3
Production Runs

3.0   PRODUCTION RUNS

3.1   Specifications

      A production run represents a "group" or "batch" of samples that are
      processed in a continuous sequence under relatively stable conditions.
      Specifically:

3.1.1    Calibration - All samples in a run use the same initial calibration
         data.  For mercury analyses, samples prepared by a certain method
         must be analyzed with calibration and QC standards prepared by the
         same method.  Therefore, all samples, calibration standards, and QC
         standards in a run must be associated with the same Preparation Code
          (Type 21 record).

3.1.2    Method number - Constant throughout a run.

3.1.3    Instrument conditions - Constant throughout a run.  Results obtained
         on different instruments cannot be combined in one run.

3.1.4    Thus, each separate group of analyses on each instrument will consist
         of a separate production run, and must be reported in a separate
         file.

3.1.5    The run numbers in a Sample Delivery Group (SDG) must be unique; that
         is, there shall only be one Run Number "1", only one Run Number "2",
         etc. in an SDG.

3.1.6    In addition, later runs within a method for an analyte shall have a
         higher run number than earlier ones.  For example, if arsenic is
         quantitated by the Inductively Coupled Plasma - Atomic Emission
         Spectroscopy  (ICP-AES) method on 01/01/1999 beginning at 12:02 and
         arsenic is later quantitated by the ICP-AES method on 01/01/1999
         beginning at 18:06, then the run beginning at 12:02 shall have a
         lower run number than the run beginning at 18:06.

3.2   Example

      The following is an example of the sequence of record types in a
      production run.

      10   Contains run header information.   Occurs once  per run.

      16   Contains additional run header information.   Occurs  once  per run.

      20   Acts as a header for the following method and instrument  parameters
           information.   Occurs at least once per run with EPA sample number
           equal to "MDL".   Analysis year,  analysis month,  analysis  day equal
           the year,  month and day the Method Detection Limit (MDL)  was
           computed.   Analyte count equals the number of the Type 30 records
           that follow.

               21   Contains the Preparation Code (field 15) and the
                    Preparation Date (fields #8,  9, 10) for the MDL.  Occurs
                    at least once per run with each Type 21 record preceded by
                    the relevant Type 20 record and immediately followed by
                    its related Type 30 record(s).

               30   Contains the Analyte Identifier "C" (field #2),  the
                    Analyte CAS Number (field #3), the MDL Label "U" (field
                    #20), and the MDL (field #21).  Occurs once for each
                    analyte used in the run.


ILM05.2                               H-8

-------
                                                   Exhibit H — Section 3
                                                  Production Runs (Con't)
20

         21

         30

20   Acts as  a header  for the  following  instrument parameter
     information.  Occurs once per  run with EPA  sample  number equal  to
     "LRV".   Analysis  year,  analysis month, analysis  day  equal the year,
     month  and day the linear  ranges were  computed.   Analyte  count
     equals the  number of Type 30,  32 and  34  groups that  follow.

         30  Contains only the  Analyte  CAS Number  and the Analyte
             Identifier.   Occurs once  for each  analyte used in the  run.

         32  Contains integration time  information for the preceding
             analyte  on the Type 30  record.

         34  Contains the  Contract Required  Quantitation Limit (CRQL)
             and Linear Range information for the  preceding analyte on
             the Type 30 record.  There are  as  many consecutive Type 34
             records  as there are different  wavelengths or masses used
             for the  analyte identified on preceding Type 30.

         30

         32

         34

20   Acts  as  a header  for the  following  instrument  parameter
     information.  Occurs once per run  with EPA  sample number equal  to
     "BCD".   Analysis  year,  analysis  month, analysis  day equal the year,
     month and day the background correction  factors  were computed.
     Analyte  count equals the number  of  the Type 30 and 35 groups that
     follow.

         30   Contains only the Analyte CAS Number  and the Analyte
              Identifier.  Occurs once for each analyte used in the run.

         35   Contains the  background and interelement correction
              information for the preceding analyte on the Type 30
              record.   There are as many consecutive Type 35 records as
              there  are interelement correction factors for the analyte
              identified on preceding Type 30.
 20

 21


 22
     30

     35

Contains header information for sample and QC data.

Contains additional information for analytical and instrument QC
samples.  Will always be preceded by a Type 20 record.
Contains additional information for analytical samples.
usually follow a Type 21 record.
Will
          30    Contains the sample level concentration,  true or added
                value and QC value for each analyte.   Occurs once for
                                 H-9
                                                              ILM05.2

-------
Exhibit H — Section 3
Production Runs (Con't)

                      each  analytical  result  for  the  EPA  sample number of  the
                      previous  Type  20 record.

               31     Reports any instrumental data necessary  to obtain  the
                      result reported  on  the  previous Type  30  record.  Will
                      always be preceded  by a Type 30 or  31 record.  For
                      Inductively Couple  Plasma - Mass Spectrometry  (ICP-MS),
                      there are as many Type  31 records as  there are isotopes
                      for the analyte  identified  on the preceding Type 30
                      record.

               30     Values for the next analyte being measured.

               31     Values for the next analyte being measured.

               30

               31  •

           Type 30-31 record sequence  continues  as many times  as  the value of
           the ANALYTE COUNT on the previous  Type 20  record.

      20   Next Sample Header  record - The following  applies to the next
           sample  data.

      21

      22

               30

               31

               30

               31  etc.

      20

      21

      22

               30

               31  etc.
ILM05.2                               H-10

-------
                                                         Exhibit H — Section 4
                                                                Record Sequence

4.0   RECORD SEQUENCE

4.1   Specifications

      A Run Header  (Type 10) record must be present as the first record in the
      file  (run).   Further occurrences of the Type 10 record in the file are
      not allowed.

4.1.1    A Type 16 record must immediately follow the Type 10 record.  Further
         occurrences of the Type 16 record in the file are not allowed.

4.1.2    The first Type 20 records with EPA sample numbers MDL, LRV, and BCD
         are headers for the run-wide method and instrument parameters.

4.1.3    The first Type 20 record of the Type 21, 30 group is a header for the
         annually determined Method Detection Limits  (MDLs) and must
         immediately follow the Type 16 record.  A Type 20 record of the Type
         21, 30 group must be present for each MDL reported in the run.  For
         ICP-AES,  ICP-MS, and cyanide analyses, an MDL associated with
         Preparation Code "NP1" must be present in each run.  This MDL shall
         be used in the qualification of the data reported for non-prepared
         samples and instrument QC analyses  (except the distilled Initial
         Calibration Verification  (ICV) standard for cyanide).

4.1.4    The next Type 20 record of the Type 30, 32, 34 group is a header for
         the Linear Range Values  (LRVs) and must immediately  follow the last
         Type 30 record of the Type 21, 30 group that pertains to the MDL.
         The linear range values for all methods except the Inductively
         Coupled Plasma - Atomic Emission Spectroscopy  (ICP-AES) and
         Inductively Coupled Plasma - Mass Spectrometry  (ICP-MS) methods are
         the analytically determined concentrations of the highest instrument
         calibration standards that are used in the generation of the
         calibration curve at the beginning of every  run.  The linear range
         values for the ICP-AES and ICP-MS methods are the quarterly
         determined values that are reported on Form XI-IN of the hardcopy.

4.1.5    The next Type 20 record of the Type 30, 35 group is a header for the
         ICP-AES Background Correction Data  (BCD) and must immediately follow
         the last Type 34 record of the Type 30, 32,  34 group that pertains to
         the linear range values.  This Type 20 record is not required for
         methods MS, AV, CV, CA, AS and C  (i.e., ICP-MS, mercury, and cyanide
         analyses).

4.1.6    These  are  the only occurrences of the Type 20 records that do not
         relate to  actual analyses in the run.  Therefore, the only  fields
         that are not blank in these occurrences of the Type  20  record are the
         RECORD TYPE  ("20"); EPA SAMPLE NUMBER  ("MDL", "LRV"  and "BCD");
         Analysis Year/Year Computed, Analysis Month/Month Computed, Analysis
         Day/Day Computed  ("YYYY", "MM", "DD"); and ANALYTE COUNT.

4.1.7    A  minimum  of one Type 30 record must immediately follow the Type 21
         record of  the Type 21, 30 group with EPA sample number  MDL, and the
         total  number of Type 30 records must be equivalent to the ANALYTE
         COUNT  on the Type  20 record.

4.1.8    A  minimum  of one Type 30, 32, 34 group with  EPA sample  number LRV
         must  immediately follow  the Type 20 record which  is  preceded by the
         last  Type  30 record of the final Type  21, 30 group.   The  information
         in each  Type 30, 32, 34 group must  pertain to one and only  one
         analyte.   The number of Type  30, 32,  34 groups must  be  equivalent to
         the ANALYTE COUNT  on the  Type 20 record.
                                      H-ll                               ILM05.2

-------
Exhibit H — Section 4
Record Sequence  (Con't)

4.1.9    A minimum of one Type 30, 35 group with EPA sample number BCD must
         immediately follow the Type 20 record for background correction data
         (if required).  This Type 20 is preceded by the last Type 34 record
         of the  final Type 30, 32, 34 group.  The information in each Type 30,
         35 group must pertain to one and only one analyte.  The number of
         Type 30, 35 groups must be equivalent to the ANALYTE COUNT on the
         Type 20 record.

4.1.10   The Type 20 record that relates to the analysis of the first
         instrument calibration standard must immediately follow the last Type
         30, 35  group for ICP-AES, or the last Type 30, 32, 34 group for
         mercury and cyanide analyses.  For ICP-MS, the Type 20 record for the
         first instrument tune standard analysis must immediately follow the
         last Type 30, 32, 34 group and the Type 20 record for the first
         instrument calibration standard must immediately follow the last 30,
         31 group from the last tune standard analyzed.  After the appearance
         of these Type 20 records in the file, further occurrences of the Type
         32, 34  and 35 records in that file are not allowed.

4.1.11   Each environmental sample, calibration, or Quality Control (QC)
         sample  is represented by a group composed of Type 20, 21, and 22
         records, which hold sample level identifying information, followed by
         a minimum of one group composed of Type 30 and 31 records for each
         analyte.  The Type 20 record holds a count for the number of analytes
         being used to determine results.  The ANALYTE COUNTER must have a
         value equivalent to the number of Type 30 groups associated with each
         Type 20 record.

4.1.12   Except  for the first Type 20 records (EPA sample numbers MDL, LRV,
         BCD) for method ICP-AES and the first two Type 20 records  (EPA sample
         numbers MDL, LRV) for the methods for ICP-MS, mercury and cyanide
         analyses, all Type 20 records should occur in the order of sample
         analysis.

4.1.13   Type 90 comment records may be defined to occupy any position except
         before  the Type 10  (header) record.  Comments pertaining to the whole
         run such as ones on Cover Page must appear before the first Type 20
         record.  Comments pertaining to a particular sample such as ones on
         Forms IA-IN and IB-IN must appear after the Type 20 record for that
         sample, but before the first Type 30 record associated with that
         sample.  Comments pertaining to a particular analyte must appear
         after the Type 30 record of that analyte, but before the Type 30
         record  of the following analyte.

4.1.14   The Type 92 record which contains the sample associated data that is
         reported at the bottom of Forms IA-IN and IB-IN must appear anywhere
         after the Type 22 record for that EPA Field Sample, but before the
         Type 20 record of the next sample.
ILM05.2                               H-12

-------
                                                      Exhibit H — Sections 5-7
                                                          File/Record Integrity
5.0   FILE/RECORD INTEGRITY
      All record types must contain the following check fields to ensure file.
      and record integrity:
                                                        Remarks
                                                        "10"  or  as appropriate
       Record         Field   Field
       Position       Length  Contents
       First Field      2     Record type or
                              identifier
       Last Field.       5     Record sequence number   00000-99999,  repeated
                                                       as necessary
                        4     Record checksum1         Four hexadecimal
                                                       digits
                        2     Must contain CR and  LF

6.0   DATES AND TIMES

      Date or time-of-day information consists of successive groups of digits/
      each separated by delimiters.   Dates are given in the order YYYY MM DD,
      and times as HH MM.   All hours must be given as 00 to 23 using a 24 hour
      clock and must be local time.   All days shall be given as 01 to 31.  All
      months shall be given as 01 to 12 (e.g., 01 is January, 02 is February).

7.0   MULTIPLE VOLUME DATA

      There is no requirement under this format that all the data from an
      entire Sample Delivery Group (SDG) fit onto a single diskette.  However,
      each single production run must fit onto a single diskette if possible.
      If that is not possible, then it is necessary that all files start with
      a Type 10 record, and that the multiple Type 10 records for each file of
      the same production run be identical.  Information for a single sample
      may not be split between files.
       lThe  checksum  is  the  sum  of  the ASCII  representation  of the  data  on  the
 record up to the Record Sequence Number (not including the  Record Sequence
 Number),  plus the checksum of the previous record.   The sum is taken modulo
 65536 (216) and is represented as four hexadecimal digits (i.e., the remainder
 of the sum divided by 65536 represented as four hexadecimal digits).
                                       H-13
                                                                         ILM05.2

-------
Exhibit H — Section 8
Deliverable

8.0   DELIVERABLE

8.1   Requirements

      The file shall be submitted on IBM-compatible, 3.5 inch, high density
      1.44 MB diskettes.  The diskettes shall be formatted and recorded using
      DOS/Windows Operating Systems.  The diskettes shall contain all
      information relevant to one and only one Sample Delivery Group  (SDG).
      An alternative means of electronic transmission may be utilized if
      approved in advance by USEPA.

8.1.1    USEPA Agency Standard data from an entire SDG may not fit onto a
         single diskette.  If a single production run is being split onto
         multiple diskettes, then all files shall start with a Type 10 record,
         and the multiple Type 10 records for each file of the same production
         run shall be identical.  Do not split the data from a single sample
         onto multiple diskettes.

8.1.2    Information on the diskette must correspond to information submitted
         in the hardcopy raw data package and on the hardcopy raw data package
         forms.  Unused records shall not be included on the diskettes.   If
         the information submitted in the hardcopy data package forms is
         changed, the information in the electronic file  (e.g., diskette)
         shall be changed accordingly, and a complete electronic deliverable
         containing all the information for the SDG shall be resubmitted  along
         with the hardcopy at no additional cost to USEPA.

8.1.3    Each diskette shall be identified with an external label containing
          (in this order) the following information:

         Disk Density
         File Name(s)
         Laboratory Name  (optional)
         Laboratory Code
         Contract Number
         Case Number/SDG
         NRAS Number  (where applicable)
         Initial Submission or Resubmission  (as applicable) and Date

8.1.4    The format for File Name shall be XXXXXX.I01 to XXXXXX.I99, where
         XXXXXX is the SDG identifier, I designates inorganics, and 01 through
         99 is the file number.

8.1.5    Dimensions of the label must be in the range of 2-1/2" to 2-3/4"  long
         by 2" to 2-1/8" wide for a 3-1/2" diskette.
 ILM05.2                               H-14

-------
                                                         Exhibit H — Section 9
                                                                 Record Listing
9.0
RECORD LISTING
      The following section provides information for the usage of each of the,
      record types.  Where specified, labels indicate the nature of the
      value(s)  that follow on that record.  If the value(s) will not be
      reported, the label shall be omitted. Listed below is every record type
      required to report data from a single Sample Delivery Group (SDG).

9.1   Production Run First Header Record  (Type 10)
      Use: Each production run will start with a Record Type 10.
         MAXIMUM LENGTH
                2
                1
                4
                1
                2
                1
                2
                1
                2
                1
                2
                1
                5
                1
                8
                1
                3
                1
                6
                4
                11
                1
                10
                2
                25
                1
                2
                1
                5
                4
                     CONTENTS
                     RECORD TYPE
                     Delimiter
                     ANALYSIS START YEAR
                     Delimiter
                     ANALYSIS START MONTH
                     Delimiter
                     ANALYSIS START DAY
                     Delimiter
                     ANALYSIS START HOUR
                     Delimiter
                     ANALYSIS START MINUTE
                     Delimiter
                     METHOD TYPE
                     Delimiter
                     METHOD NUMBER
                     Delimiter
                     MANAGER'S  INITIALS
                     Delimiter
                     LAB  CODE
                     Delimiter
                     CONTRACT NUMBER
                     Delimiter
                     INSTRUMENT ID
                     Delimiter
                     LABORATORY NAME
                     Delimiter
                     RUN  NUMBER
                     Delimiter
                     RECORD  SEQUENCE  NUMBER
                     CHECKSUM
FORMAT/CONTENTS
"10"
I
YYYY
I
MM
I
DD
I
HH
I
MM
I
CHARACTER 2
I
"ILM05.2"  (SOW)
I
CHARACTER
I
CHARACTER
I I I I
CHARACTER
I
CHARACTER
I I
CHARACTER
 I
NUMERIC 3
 I
NUMERIC
CHARACTER
       2Analysis Method Types are:
       "P" for ICP-AES
       "MS" for ICP-MS
       "CV" for Manual Cold Vapor AA
       "AV" for Automated Cold  Vapor AA
       "AS" for Semi-Automated  Spectrophotometric
       "C" for Manual Spectrophotometric

       3Run number values  are 01  through  99.  Each production  run will be
 assigned a unique Run Number.   Run Numbers are to  be assigned sequentially
 beginning with 01 and will equal the number of production runs.
                                       H-15
                                                                   ILM05.2

-------
Exhibit H — Section 9
Record Listing (Con't)

9.2   Production Run Second Header Record  (Type 16)
MAXIMUM LENGTH
2
I
4
1
2
I
2
1
2
1
2
1
1
1
1
1
1
1
1
1
5
4
CONTENTS
RECORD TYPE
Delimiter
ANALYSIS END YEAR
Delimiter
ANALYSIS END MONTH
Delimiter
ANALYSIS END DAY
Delimiter
ANALYSIS END HOUR
Delimiter
ANALYSIS END MINUTE
Delimiter
AUTO-SAMPLER USED
Delimiter
INTERELEMENT CORRECTIONS APPLIED
Delimiter
BACKGROUND CORRECTIONS APPLIED
Delimiter
RAW DATA GENERATED
Delimiter
RECORD SEQUENCE NUMBER
CHECKSUM
FORMAT/CONTENTS
"16"
1
YYYY
1
MM
1
DD
1
HH
1
MM
1
"Y" or "N" 4
1
"Y" or "N" 5
1
"Y" or "N" 5
1
"Y" or "N" or "B'
1
NUMERIC
CHARACTER
      4Enter "Y" if an auto-sampler is used with equal time and intervals
between analysis.

      5These are the answers to the first two questions on the Cover Page of
the hardcopy deliverable.  "Y" equals "YES", and "N" equals "NO".

      6This is the answer to the third question on the Cover Page of the
hardcopy deliverable.  "Y" equals "YES", "B" equals BLANK, and "N" equals
"NO".
ILM05.2
                                      H-16

-------
9.3   Mandatory Sample Header Data Record (Type 20)
  MAXIMUM LENGTH
         2
         1
         2
         1
         12
         1
         5
         1
         3
         1
         3
         1
         5
         1
         6
         1
         4
         1
         2
         1
         2
         1
         2
         1
         2
         2
         2
         1
         5
         1
         3
         1
         5
         4
CONTENTS
RECORD
Delimiter
REGION
Delimiter
EPA SAMPLE NUMBER
Delimiter
MATRIX
Delimiter
QC CODE
Delimiter
SAMPLE QUALIFIER
Delimiter
CASE NUMBER
Delimiter
SDG NUMBER
Delimiter
ANALYSIS YEAR/YEAR COMPUTED
Delimiter
ANALYSIS MONTH/MONTH COMPUTED
Delimiter
ANALYSIS DAY/DAY COMPUTED
Delimiter
ANALYSIS HOUR
Delimiter
ANALYSIS MINUTE
Delimiter
SAMPLE WT/VOL UNITS
Delimiter
SAMPLE WT/VOL
Delimiter
ANALYTE COUNT
Delimiter
RECORD SEQUENCE NUMBER
CHECKSUM
                                                         Exhibit H — Section 9
                                                         Record Listing (Con't)
 FORMAT/CONTENTS
 "20"
 I
 NUMERIC
 I
CHARACTER 7
CHARACTER8
I
CHARACTER
I
CHARACTER
I
CHARACTER
I
CHARACTER
I
YYYY
I
MM
 I
DD
 I
HH
 I
MM
 II
"G"/"ML"'
 I
NUMERIC10
 I
NUMERIC
 I
NUMERIC
CHARACTER
       7EPA Sample Number as it appears on Form XIII-IN of the hardcopy
 deliverable except for the first Type 20 records.  The first Type 20 record
 must have an EPA sample number of  "MDL"; after all Type 20  records with an EPA
 sample number of "MDL", the next Type 20 record must have an EPA sample number
 of  "LRV"; for ICP-AES, the Type 20 record following the "LRV" must have an EPA
 sample number of "BCD".

       "For matrix,  "1" equals "WATER" and "F" equals "SOIL".  A matrix
 identifier  ("1" or "F") is required  for all EPA sample numbers except "BCD".

       9"G" equals grams and "ML" equals milliliters.

       10This is  the size of the sample at  the  beginning of the digestion
 procedure.
                                      H-17
                                                     ILM05.2

-------
Exhibit H — Section 9
Record Listing  (Con't)

9.3.1    SAMPLE QC CODES LISTING FOR TYPE 20

         NOTE: These QC codes appear in  the QC code  field on  the  Type  20
         record  (R20F5).  They are used  to indicate  the  type  of data that  is
         being reported.
   QCC     Name

   LRB     LABORATORY (REAGENT)
           BLANK

   LCB     LABORATORY CALIBRATION BLANK
   LIB     LABORATORY INITIAL BLANK
 Definition

 The  Preparation  Blank (see
 Exhibit  G).

The Continuing Calibration Blank
 (CCB)  (see Exhibit  G).

The Initial Calibration Blank
 (ICB)  (see Exhibit  G).
   LCM     LABORATORY CONTROL
           SOLUTION
 The  Laboratory  Control Sample
 (LCS)  (see Exhibit G).
   LD2     LABORATORY DUPLICATE
           SECOND MEMBER
 This  is  the  second aliquot and
 is  identified as  "D"  on Form VI-
 IN  of the  hardcopy.
   LVM     LABORATORY CALIBRATION
           VERIFICATION SOLUTION
   LVC     LABORATORY CONTINUING
           CALIBRATION VERIFICATION
   LVD     LABORATORY DISTILLED
           VERIFICATION SOLUTION
 These values  are identified as
 "Initial  Calibration
 Verification" (ICV)  on Form IIA-
 IN of hardcopy.

 These values  are identified as
 "Continuing Calibration
 Verification" (CCV)  on Form IIA-
 IN of hardcopy.

 These values  are the "distilled
 ICV"  results  for cyanide.   Refer
 to Exhibit D,  Section 12.7.1 for
 cyanide.
   LSF     LABORATORY SPIKED SAMPLE -
           FINAL VALUES
 These are the "Spiked Sample
 Result (SSR)" values of Form VA-
 IN of hardcopy.
   LDO     LABORATORY DILUTED SAMPLE
           BACKGROUND (ORIGINAL) VALUES
   LDF     LABORATORY DILUTED SAMPLE -
           FINAL VALUES
 These values are the "Initial
 Sample Result (I)" values on
 Form VIII-IN of hardcopy.

 These are the "Serial Dilution
 Result(S)" values Form VIII-IN
 of hardcopy.
ILM05.2
                                      H-18

-------
                                                      Exhibit H — Section 9
                                                      Record Listing  (Con't)
PDO    POST-DIGESTION  SPIKE
       BACKGROUND  (ORIGINAL)  VALUES
PDF     POST-DIGESTION  SPIKE
        BACKGROUND  (FINAL)  VALUES
                                 This value is identified as
                                 "Sample Result" (SR)  on Form VB-
                                 IN of hardcopy.
                                 This value is identified as
                                 "Spiked Sample Result" (SSR)
                                 Form VB-IN of hardcopy.
                                                                      on
LPC
CRQL CHECK STANDARD
LSA
LSB
LTS
LABORATORY INTERFERENCE CHECK
SOLUTION A
LABORATORY INTERFERENCE CHECK
SOLUTION AB
LABORATORY TUNE SAMPLE
Laboratory Performance Check
 Solution  for  analysis methods  P,
 MS, CV, AV, AS,  and C  (EPA
 sample  number is CRI##).   These
 results are reported on  Form
 IIB-IN  of hardcopy.

 The results of this solution
 analysis  (EPA sample number is
 ICSAlf) are reported on  Forms
 IVA and IVB-IN of hardcopy.

 The results of this solution
 analysis  (EPA sample number is
 ICSABlt)  are  reported  on Forms
 IVA and IVB-IN of hardcopy.

The results of these solution
 analyses  are  reported  on Form
 XIV-IN  of hardcopy.
 FRB
        FIELD BLANK
                                 This is any sample that is
                                 submitted  from the field and is
                                 identified as a blank.  This
                                 includes trip blanks, rinsates,
                                 equipment  blanks, etc.
 FRM    PERFORMANCE EVALUATION  (PE)
        SAMPLE
 FLD    FIELD SAMPLE
 ZZQ    NON-SDG SAMPLE
                                  This  is  a  sample  of  known
                                  composition  provided by  USEPA
                                  for Contractor analysis  and  is
                                  used  to  evaluate  Contractor
                                  performance.

                                  This  is  the sample that  is
                                  identified by a unique EPA
                                  sample number on  the Traffic
                                  Report/Chain of Custody  Record.

                                  This  is  any sample that  is
                                  analyzed and is not  part of  the
                                  SDG  (EPA sample number is
                                  ZZZZZZ).
 STB
 CALIBRATION STANDARD
 This  is the instrument
 calibration Blank Standard
 sample number is SO).
                                                                     (EPA
                                    H-19
                                                               ILM05.2

-------
Exhibit H — Section 9
Record Listing  (Con't)
   STC     CALIBRATION STANDARD
   STD     CALIBRATION STANDARD
                                This is the instrument
                                 calibration CRQL Standard (EPA
                                 sample number is Sx where x is
                                 the CRQL value of the analyte).

                                This is the instrument
                                 calibration standard other than
                                 the Blank Standard or the CRQL
                                 Standard (EPA sample number is
                                 S) .
   STM
           MIORANGE STANDARD
   STR
           RESLOPE SAMPLE
   STL
BASELINE SAMPLE
This is the distilled cyanide
Mid-range  Standard  (EPA  sample
number  is  MIDRANGE##).   Refer to
Exhibit D,  Section  10.2.1.1, for
cyanide.

This is the resloping that is
permitted  for mercury analysis
 (EPA sample number  is
RESLOPE**). Refer to  Exhibit D,
Section 9.1.5,  for  mercury.

This is the baseline correction
that is permitted for mercury
analysis  (EPA sample  number is  .
BASELINE**).  Refer to Exhibit
D,  Section 9.1.5, for mercury.
   MDQ
   LRQ
           METHOD DETECTION LIMIT
LINEAR RANGE VALUE
   BCQ
           BACKGROUND CORRECTION
These are the annually
 determined  analyte  detection
 limits  that are  reported on  Form
 IX-IN of  hardcopy.   (EPA sample
 number  is MDL).

These are the quarterly
 determined  values for ICP-AES
 and ICP-MS  methods  that  are
 reported  on Form XI-IN of
 hardcopy.   For all  other
 methods,  these are  the
 analytically determined
 concentrations of the highest
 instrument  calibration standards
 that are  used in the generation
 of the  calibration  curve at  the
 beginning of every  run.   (EPA
 sample  number is LRV).

These are the ICP-AES  annually
 determined  interelement
 correction  factors  that  are
 reported  on Forms XA and XB-IN
 of hardcopy.  (EPA  sample number
 is BCD).
         NOTE: All  field samples  that  are  reported on  the  Traffic  Report/Chain
         of  Custody Record  shall  contain the  QC  code "FLD" in  Record Type  20
ILM05.2
                                      H-20

-------
                                                Exhibit H — Section 9
                                                Record Listing  (Con't)

Field Number 5 (R20F5) except when "FLD" is superseded by "FRB"
(Field Blank Sample),  "FRM" (PE Sample).

For Matrix Spike and Duplicate sample analysis  (Forms VA-IN and VI-IN
of hardcopy), the "Sample" result shall contain the QC code "FLD" in
R20F5, the "Spiked Sample Result" shall contain the QC Code "LSF" in
R20F5, and the "Duplicate" result shall contain the QC code "LD2" in
R20F5.
                              H-21                               ILM05.2

-------
Exhibit H — Section 9
Record Listing (Con't)

9.4   Sample Header Record  (Type 21)
  MAXIMUM LENGTH
         2
         2
         3
         2
         3
         1
         6
         1
         14
         1
         4
         1
         2
         1
         2
         2
         4
         1
         2
CONTENTS
RECORD TYPE
Delimiter
LEVEL
Delimiter
PREPARATION CODE
Delimiter
NRAS NUMBER
Delimiter
LAB SAMPLE ID
Delimiter
PREPARATION YEAR
Delimiter
PREPARATION MONTH
Delimiter
PREPARATION DAY
Delimiter
YEAR RECEIVED
Delimiter
MONTH RECEIVED
FORMAT/CONTENTS
"21"
I I
"LOW"/"MED"
I I
CHARACTER11
I
CHARACTER
I
CHARACTER
I
YYYY
MM
 I
DD
 I I
YYYY
 I
MM
      "Preparation Codes:  A Preparation Code  is  required for  all  EPA  sample
numbers except "LRV", "BCD", and "TUNE##".
"HW1" -  Hotplate/Block digestion  for ICP-AES analysis of water samples.
"HW2" -  Hotplate/Block digestion  for ICP-MS analysis of water samples.
"MW1" -  Microwave digestion for ICP-AES analysis of water samples.
"MW2" -  Microwave^digestion for ICP-AES analysis of water samples.
"HS1" -  Hotplate/Block digestion  for ICP-AES analysis of soil samples.
"HS2" -  Hotplate/Block digestion  for ICP-AES analysis of soil samples.
"MSI" -  Microwave digestion for ICP-AES analysis of soil samples.
"CW1" -  Preparation for the Manual Cold Vapor AA analysis of water samples.
"CS1" -  Preparation for the Manual Cold Vapor AA analysis of soil samples.
"CW2" -  Preparation for the Automated  Cold Vapor analysis of water samples.
"DW1" -  Distillation for  the manual and semi-automated  spectrophotometric
         analysis  of water samples.
"DW2" -  Midi-distillation for  the semi-automated spectrophotometric  analysis
         of  water  samples.
"DS1" -  Distillation for  the manual and semi-automated  spectrophotometric
         analysis  of soil  samples.
"DS2" -  Midi-distillation for  the semi-automated spectrophotometric  analysis
         of  soil samples.
"NP1" -  No  preparation.
 ILM05.2
                  H-22

-------
Sample Header Record  (Type 21)  
-------
Exhibit H — Section 9
Record Listing  (Con't)

9.5   Associated Injection and Counter Record  (Type 22)

  MAXIMUM  LENGTH    CONTENTS                             FORMAT/CONTENTS
         2          RECORD TYPE                          "22"
         8          Delimiter                            I I I I I I I I
         5          VOLUME ADJUSTMENT FACTOR           NUMERIC15
         2          Delimiter                            I |
         8          FINAL VOLUME                         NUMERIC16
         1          Delimiter                            I
         8          DILUTION FACTOR                     NUMERIC
         3          Delimiter                            I | |
         5          PERCENT SOLIDS                       NUMERIC
         1          Delimiter                            I
         5          RECORD SEQUENCE NUMBER             NUMERIC
         4          CHECKSUM                             CHARACTER
      15This field is used to report any additional volume adjustments in the
preparation method.   As  an  example,  the factor  of  1.25  that  results  from the
chloride interference volume adjustment in  Preparation  Method/Code HW2.

      16This is the final volume that is currently reported on Form XII-IN of
the hardcopy.

ILM05.2                               H-24

-------
9.6   Results Data Record  (Type 30)
  MAXIMUM LENGTH
         2
         1
         1
         1
         9
         2
         5
         1
         3
         1
         15
         1
         1
         1
         10
         1
         1
         1
         10
         1
         1
         1
         10
         1
         1
         1
         10
         1
         10
         1
         1
         1
         1
         1
         10
         2
         15
         1
         10
         1
         5
         4
CONTENTS
RECORD TYPE
Delimiter
ANALYTE IDENTIFIER
Delimiter
ANALYTE CAS NUMBER
Delimiter
CONCENTRATION UNITS
Delimiter
CONCENTRATION QUALIFIER
Delimiter
CONCENTRATION
Delimiter
VALUE DESCRIPTOR
Delimiter
AMOUNT ADDED OR TRUE VALUE
Delimiter-
QC VALUE DESCRIPTOR, P
Delimiter
QC VALUE
Delimiter
QC VALUE DESCRIPTOR, L
Delimiter
QC VALUE
Delimiter
MATRIX SPIKE QC LIMIT QUALIFIER
Delimiter
QC LOWER LIMIT
Delimiter
QC UPPER LIMIT
Delimiter
QC LIMIT QUALIFIER
Delimiter
MDL  LABEL
Delimiter
MDL
Delimiter
RAW  DATA AVERAGE
Delimiter
RAW  DATA %RSD
Delimiter
RECORD SEQUENCE NO.
CHECKSUM
                                                          Exhibit H — Section 9
                                                          Record Listing (Con't)
 FORMAT/CONTENTS
 "30"
CHARACTER18
 1 1
CHARACTER19
 I
NUMERIC20'21
 I
Ť m n I \\ pi // 2 2

 I
NUMERIC
NUMERIC
 I I I
 NUMERIC
 I

 I
 NUMERIC "
 I
 NUMERIC "
 I
 \\ it tr / \\Tjirr26

 I
 "U"
 I
 NUMERIC 27
 I I
 NUMERIC28
 I I
 NUMERIC
 I I
 NUMERIC
 CHARACTER
                                       H-25
                                                      ILM05.2

-------
Exhibit H — Section 9
Record Listing  (Con't)

             FORMAT OF THE RESULTS DATA RECORD (TYPE 30) FOOTNOTES

      17"C"  (CAS Registry Number)  is  used for all  metals and cyanide.

      18The  CAS  Numbers  for metals and cyanide are in Exhibit B,  Form  IA-IN,
and Table 1 - Inorganic Target Analyte List and Contract Required Quantitation
Limits  (CRQLs), in Exhibit C.  NOTE: The CAS Numbers for the  ICS non-target
interferents are as follows: carbon  (7440-44-0);  chlorine  (7782-50-5);
molybdenum  (7439-98-7); phosphorus  (7723-14-0); sulfur  (7704-34-9), and
titanium  (7440-32-6).

      19"BDL" means  below detection limit.

"NSQ" means there is not  sufficient  quantity  to prepare sample according
specification in Exhibit  D; therefore, a smaller  sample size  is used.

"NAR" means no  analysis result required.

"LTC" means less than the CRQL but greater than or equal to the MDL.

"FQC" means failed Quality Control  (QC) criteria.

"GTL" means greater than  the linear  range.  The result  is  reported from a re-
analysis at an  appropriate dilution.

"RIN" means that the analysis result was not  used to report data in the SDG.
The result  is reported from a later  re-analysis of the  same sample aliquot.
"REX" means that the analysis result was not  used to report data in the SDG.
The result  is reported from a later  re-analysis of a repreparation of the same
sample.

Note that,  except for WNAR", none of these codes  relieves  the Contractor from
reporting a valid result.  They only explain  why  or  if  the result is
qualified.

      20EPA  Field Samples reported on Traffic Report/Chain  of Custody  Record
(QC codes FLD,  FRB, FRM)  shall have  their analytes'  results reported to four
decimal places.

     ^ 21Follow  the  instructions  for the reporting  of  data in Exhibit B in
reporting results for samples with QC  codes.   For example, the LD2 QC code
sample results  shall be reported  to  four decimal  places because the duplicate
results on  Form VI-IN have to be  reported to  four decimal  places.  Refer to
Section 9.3.1 for QC codes and definitions.

      22"T"  stands  for an analyte's true value in  a solution.   This  includes
the concentration of all  Instrument  Calibration Standards  for ALL methods of
analysis.   "F"  stands for an added concentration  to  a sample  such as a pre- or
post-digestion  spike.

      23Ťpť  equals  Percent Recovery  (%R),  Percent  Difference  (%D),  Relative
Percent Difference  (RPD),  Percent Relative Standard  Deviation (%RSD), Percent
Relative  Intensity  (%RI),  or correlation coefficient.   "L" equals control
limit for duplicates.  The matrix spike  sample %R shall be entered on the Type
30 record of the EPA sample number with  the  "S" suffix  (QC code=LSF).  The
post digest spike sample  %R shall be entered on the  T.ype 30  record of the EPA
sample number with  the "A"  suffix (QC  code=PDF).  The RPD  and the control
limit for duplicates shall be entered  on the  Type 30 record of the EPA sample
number with the "D" suffix  (QC code=LD2).  The ICP serial  dilutions %D shall
be entered  on the Type 30 record  of  the EPA  sample number  with the "L" suffix
(QC code=LDF).   The average %RSD  for ICP-MS  tune  analyses  shall be entered  on
the Type  30 record  of the last EPA sample number  "TUNE#f"  (QC code=LTS) in
each run.   The  %RI  for ICP-MS internal standards  shall  be  entered on the Type
30 record of all EPA samples numbers (except  "TUNEf#",  "ZZZZZZ", "MDL", and
"LRV").   The correlation  coefficient for the  calibration for  mercury and


ILM05.2                                H-26

-------
                                                         Exhibit H — Section 9
                                                         Record Listing  (Con't)

cyanide analyses shall be reported on the Type 30 record of the EPA sample
number associated with the final standard analyzed in the calibration curve
(immediately preceding the ICV).

      2<"N"  is the qualifier that is used on Form VA-IN of the hardcopy to
indicate that the matrix or pre-digestion spike sample recovery for an analyte
is not within the specified control limits.  The "N" qualifier shall be
entered on the Type 30 record of the EPA sample number with the WS" suffix  (QC
code=LSF).

      "These are the  control limits for the ICV/CCV percent  recovery  (%R)  on
Form IIA-IN, the CRI %R on Form IIB-IN, the ICSA/ICSAB %R on Forms IVA and
IVB-IN, the matrix spike %R on Form VA-IN, and the LCSW %R and the LCSS upper
and lower limits on Form VII-IN.  The QC upper and lower limits for the Spike
Sample Recovery shall be entered on the Type 30 record of the EPA sample
number with the "S" suffix (QC code=LSF).

      26Ť*Ť  is tne qualifier that is used  on Form VI-IN of the hardcopy to
indicate that the duplicate sample analysis for an analyte is out of control,
and "E" is the qualifier that is used on Form VIII-IN of the hardcopy to
indicate that the ICP serial dilution analysis results are estimated because
of the existence of significant physical or chemical interferences.  The "*"
qualifier should be entered on the Type 30 record of the EPA sample number
with the "D" suffix  (QC code=LD2) The "E" qualifier shall be entered on the
Type 30 record of the EPA sample number with the "L" suffix  (QC code-LDF).

      "The  MDL  shall  be  reported to  2  significant  figures for values  less
than 10 and to 3 significant figures for values greater than or equal to 10.
MDLs shall be reported in UG/L for water samples, ICV, ICB, CCV, CCB, CRI,
ICSA, ICSAB and MIDRANGE  (for cyanide), and any other samples with
concentration results reported in "UG/L".  MDLs shall be reported in MG/KG  for
soil samples.

      28The  average  value of  the replicate injections  or  exposures  are
reported  in this field.  The average values for mercury and cyanide analyses
are also  reported in  this field.  In addition, the raw data average value
shall always be reported in units of UG/L  to a minimum of four decimal places,
regardless  of the units the instrument  readings are reported in, on record
Type 31.  The raw data average value shall not be corrected  for dilutions or
volume adjustments.

For Instrument Calibration Standards analyses and Instrument Tune Standards
analyses, the raw data average  is not  required to be  reported.
                                       H-27                               ILM05.2

-------
Exhibit H — Section 9
Record Listing  (Con't)

9.7  . Instrumental Data Readout  (Type  31)

    MAXIMUM LENGTH      CONTENTS                       FORMAT/CONTENTS
           2            RECORD TYPE                    "31"
           1            Delimiter                      I
           1            TYPE OF DATA                   "W"/"M"29
           1            Delimiter                      I
           1            TYPE OF VALUE                  CHARACTER30
           2            Delimiter                      I|
           8            ANALYTE WAVELENGTH/MASS        NUMERIC (TO 2  DECIMAL PLACES)
           1            Delimiter                      I
          15            FIRST INSTRUMENT VALUE         NUMERIC 31
           2            Delimiter                      I |
          15            SECOND INSTRUMENT VALUE        NUMERIC 31
           2            Delimiter                      I |
          15            THIRD INSTRUMENT VALUE         NUMERIC 3l
           2            Delimiter                      I|
          15            FOURTH INSTRUMENT VALUE        NUMERIC 31
           2            Delimiter                      I|
          15            FIFTH INSTRUMENT VALUE         NUMERIC 31
           1            Delimiter                      I
           5            RECORD SEQUENCE NUMBER         NUMERIC
           4            CHECKSUM                       CHARACTER
      29"
        "W"  equals  wavelength,  "M" equals mass.
      30"C"  equals  concentration in ug/L,  "B" equals  absorbance,  "I" equals
intensity (counts per second or  equivalent) .

      31Used to  report data for method analyses that  require replicate
injections or exposures.   If a  single  instrument measurement is used, then
enter it in the first instrument value field,  and leave the other four fields
empty.  If two instrument  measurements are  used, then enter them in  the first
and second instrument value fields  in  the order of their analyses,  and leave
the other three fields empty, etc.   In addition, the instrument values shall
be reported to a minimum of four decimal places.

ILM05.2                                H-28

-------
9.8   Auxiliary Data Record (Type 32)
    MAXIMUM LENGTH
           2
          10
           2
           1
          10
           4
           5
           4
CONTENTS
RECORD TYPE
Delimiter
INTEGRATION TIME CODE
Delimiter
INTEGRATION TIME
Delimiter
RECORD SEQUENCE NUMBER
CHECKSUM
                                                         Exhibit H —  Section  9
                                                         Record Listing  (Con't)
 FORMAT/CONTENTS
 "32"
 I I I I I II III
WT m tt

 I
IN SECONDS
 I I I I
NUMERIC
 CHARACTER
                                       H-29
                                                  ILM05.2

-------
Exhibit H — Section 9
Record Listing  (Con't)

9.9   QC Limit Record  (Type 34)
  MAXIMUM LENGTH
         2
         4
         8

         I
         10
         1
         10
         6
         5
         4
CONTENTS
RECORD TYPE
Delimiter
ANALYTE WAVELENGTH OR MASS

Delimiter
CRQL
Delimiter
LINEAR RANGE VALUE
Delimiter
RECORD SEQUENCE NO.
CHECKSUM
 FORMAT/CONTENTS
 "34"
 I I I I
NUMERIC (TO 2 DECIMAL
 PLACES)
 I
 NUMERIC
 I
NUMERIC
 I I I I I I
NUMERIC
 CHARACTER
 ILM05.2
                                       H-30

-------
                                                         Exhibit H — Section 9
                                                         Record Listing (Con't)
9.10  Correction Data Record (Type 35)
  MAXIMUM LENGTH    CONTENTS                            FORMAT/CONTENTS
         2          RECORD TYPE                         "35"
         1          Delimiter                           I
         3          TYPE OF CORRECTION                  "ICP"
         1          Delimiter                           I I I I  I
         9          CAS NUMBER OF INTERFERING ANALYTE   CHARACTER
         1          Delimiter                           I
         8          ANALYTE WAVELENGTH                  NUMERIC (TO 2 DECIMAL
                                                        PLACES)
         1       .   Delimiter                           I
         10          CORRECTION FACTOR                   NUMERIC
         1          Delimiter                           I
         5          RECORD SEQUENCE NO.                 NUMERIC
         4          CHECKSUM                            CHARACTER
                                       H-31  •                             ILM05.2

-------
Exhibit H — Section 9
Record Listing (Con't)

9.11  Comment Record  (Type 90)

  MAXIMUM LENGTH    CONTENTS                            FORMAT/CONTENTS
         2          RECORD TYPE                         "90"
         1          Delimiter                           I
         67          ANY COMMENT                         CHARACTER
         1          Delimiter                           |
         5          RECORD SEQUENCE NUMBER              NUMERIC
         4          CHECKSUM                            CHARACTER
ILM05.2                               H-32

-------
9.12  Sample Associated Data Record (Type 92)
  MAXIMUM LENGTH
         2
         1
         9
         1
         9
         1
         6
         1
         6
         1
         6
         1
         3
         1
         5
         4
CONTENTS
RECORD TYPE
Delimiter
COLOR BEFORE
Delimiter
COLOR AFTER
Delimiter
CLARITY BEFORE
Delimiter
CLARITY AFTER
Delimiter
TEXTURE
Delimiter
ARTIFACTS
Delimiter
RECORD SEQUENCE NUMBER
CHECKSUM-
                                                         Exhibit H — Section 9
                                                         Record Listing  (Con't)
FORMAT/CONTENTS
Ť Q O "
CHARACTER
I
CHARACTER
CHARACTER
CHARACTER
CHARACTER
"YES"/BLANK
 I
NUMERIC
CHARACTER
                                       H-33
                                                     ILM05.2

-------
THIS PAGE INTENTIONALLY LEFT BLANK

-------
               APPENDIX A -- FORMAT OF RECORDS FOR SPECIFIC USES
                                   DISCLAIMER
The USEPA does not warrant or guarantee the completeness and/or accuracy of
the representative examples of record type uses provided in this appendix.
This appendix serves as an example for the usage of record types and in no way
redefines or supersedes the specifications or requirements stated in Exhibits
A through H of ILM05.2.
                                  Appendix A-l                           ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                       Appendix A-2

-------
                     Appendix A — Format of Records for Specific Uses

                                     Table of Contents

      Section                                                                     Page

1.0   ICP	5
      1.1   ICP-AES	5
      1.2   ICP-MS   	7

2.0   MERCURY	10
      2.1   Start of a Mercury Run for Water Samples with Record Types 10
            and 16 and the First Type 20 Records    	10
      2.2   Mercury Instrument Calibration Standards: Blank  (SO) and
            Four Other Standards   	10
      2.3   Spike Sample Recovery and Duplicates Performed on Different
            Samples  (QC Codes FLD, LSF, FLD, LD2)	11
      2.4   Duplicates and Spike Sample Recovery Performed on the  Same
            Sample  (QC Codes FLD, LD2, LSF )	11
      2.5   Initial Calibration Verification  (ICV)  with LVM QC Code	12
      2.6   Laboratory Control Sample  (Solid) with  LCM QC Code    	12

3.0   CYANIDE	12
      3.1   Start of a Cyanide Run with Record Types 10 and 16 and the
            First Type 20 Records	12
      3.2   Cyanide Instrument Calibration Standards: Blank  (SO) and  Five
            Other Standards	12*
      3.3   Preparation Blank  (Soil) with LRB QC Code	13
      3.4   Laboratory Control Sample  (Soil) with LCM QC Code	13
      3.5   Continuing Calibration Verification  (CCV) with LVC QC  Code    	13
      3.6   Spike Sample Recovery and Post Distillation Spike Sample  Recovery
            Performed on the Same Sample  (QC Codes  FLD, PDO, LSF,  PDF)    	 13
                                        Appendix A-3                           ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                       Appendix A-4

-------
                                                                                     Appendix A
                                                            Format of Records for Specific Uses
1.0   ICP
1.1
ICP-AES
1.1.1 Start of an ICP-AES Run with Record Types 10 and 16 and the First Type 20 Records

      10|1999|09|17|09|06|P|ILM05.2|ABC|TESLAB||||68-D2-0039|P2||TEST LABSINC.|2|000001879
      16|1999|09|17|12|03|Y|Y|Y|N|000012114
      20|1|MDL|1|MDQ|
      21| | | INP1I I I  I I I I I  I I I I
      30|C|7440-22-4|IUG/L?
      30|C|7429-90-5||UG/L|
      30ICI7440-39-3I|UG/L|
      30|C|7440-41-7||UG/L|
                  1999|07|15| |  | |
                      I000053CD5
                       I I
                      I I I
                      I I
                      I I
      20|1|MDL|1|MDQ| | | 11999 I 07
      21| I |  IHW1I| |1999|07|15| I  I
      30|C|7440-22-4| |UG/L| I I I  I
      30|C|7429-90-5| |UG/L|I I I  I
      30|C|7440-39-3| |UG/L| I I I  I
      30|C|7440-41-7| |UG/L| I I I  I

      20|1|MDL|F|MDQ||||1999)07
      21) | |  |HS1| | |1999|07|15| I  I
      30|C|7440-22-4||MG/KG||||
      30|C|7429-90-5||MG/KG||||
      30|C|7440-39-3|IMG/KGII|I
      30|C|7440-41-7||MG/KG||||
      20)1)
      30|C|
      32) | |
      34) | |
      30|C|
      32) | |
      34| | |
      30|C|
      32) | |
      34) | |
      30|C|
      32) | |
      34) | |
     LRV) 1
     7440-
     Mill
     1328.
     7429-
     HIM
     1308.
     7440-
     I  I II I
     1493.
     7440-
     Mill
     1313.
 ILRQIiii1999)07
22-4) |I | I I I
I|IT|5.00||
00 |5|40000|
90-51 I I I I I I
IIITI5.00I
201200110000001
39-31 |I I II I
IIITI5.00I|
40|20|100000|I
•41-7| MUM
 IIITI5.00II
 00121250001
I I I
    16
    II
    I I
    I I
    I I
    I I

    15
    I I
I000256CDA
    I I
I000288BB6
                        |04|000044B9D

                        I |U|3.1| Mil 1000065996
                        I|U|21.8| MMI0000767D1
                        IIUI11.5I||||I0000875CB
                        I|U|1.1|||||I0000983C5
||04|0000104B9D
00|0000113CD5
I I IUI3.4I|| || 10000125996
I I |U|22.8| | | I II00001367D1
I||U|12.5|||||I00001475CB
I||U|2.1||||II00001583C5

I|04|000164B9D
OOI00017A212
| | |IUI0.82II I I II00018C248
| | ||U|4.8| |||II00019B321
IIIIUI3.1IMM I00020CE75
|| ||U|0.42|| II II00021A21B

I|04|0002356C2
II II I I0002463D1
    000267591
            I I I I0002782AD
I00031AB1A
       I0002994FB
          I I I I I I00030A211
     I00032B436
Ml  I I I I I I I I I I I I00033C149
I00034CA52
I I I I00035D2DA
      20|1|BCD| IBCQI | ||1999|07|01| Mil I 04|0007894FB
      30|C|7440-22-4| I I I I I IM II I I I I I I II II I I I00079A20A
      35IICPI || | 17439-89-61259.90|-0.0002500 I00080AC9B
      35|ICP||||17439-96-5)257.6010.0002200|00081B6F4
      30|C|7429-90-5| I I I I I I I I I I I I I I II I II I II I00082C410
      35|ICP| | | | |7439-96-5|257.60|0.0004900 I00083CE72
      35|ICP|||||7440-62-2|292.40|-0.0419200|00084D8EF
      30ICI7440-39-3I I I I I I II I I I I IIM I I I I I I I I00085E605
      35|ICP.| | | | |7439-96-5|257. 60|0. 0000600 I 00086F060
      30|C|7440-41-7| I I I I I I II II II I I I I II I I II I00087FD73
      35|ICP|||||7440-50-8|324.70)0.0046200(000891401
      35|ICP| | | | |7439-96-5|257.60|0.0015400 I000901F30

1.1.2 ICP-AES  Instrument Calibration  Standards, SO and S

      20|1|HH|1|STB||20596|MAX123|1999|09|17|09|06||||04|00128D199
      21| | | INP1I |STDB11999|09117| | | | |TESLAB| | | I00129DD31
      22| | | | | | | | | | |1.0| | | I00130E598
                                            Appendix A-5
                                                                            ILM05.2

-------
Appendix A
Format of Records for Specific Uses (Con't)

   30 | C| 7440-22-4 | III|T|0.0||||||!||||U|3.1||||| (00131F8F5
      31|W|I| |328.00|0.0304| |0. 03741 |0. 04001
      30|C|7429-90-5| I|||T|0.0|||||||||||U|21
      31 | W|I| |308.20|0.0104| | 0.01361 10.01201 |
      30|C|7440-39-3| I I I |T|0.0| I I I I I I I I I |U|11
      31|W|I| |493. 401-0. 00021 |0. 00021 |0. 00001
      30|C|7440-41-7| | | | |T|0.0| I I I I I I I | | |U|1.1| III! I0013751FA
      31 1 W|I| 1313.0010.00061 | 0.00021 | 0.00041 | | | I001385C04
                                           | 1001320305
                                           8||||| 1001331697
                                           | 1001342137
                                           5| I | | | I00135348D
                                           | | I001363EA4
   20|1|H|1|STD| I 20596 | MAX123 1 1999 | 09 1 17 | 09 |
   21| | | INP1I |STD1|1999|09|17| | | | |TESLAB| I | (
   22| | | | | | | | | | ll.OU | I00208453C
                                               11) I I |04|00206314E
                                               002073CD5

                                                I I II 1002139157
                                                002149B6E
                                               8)1)11 (00215ADE2
                                               I00216B7EC
                                               5|| | I I I00219E77D
                                               I00220F18F
                                               I I I I I 1002210410
                                               I002220E25

1.1.3 Duplicates,  Spike Sample Recovery,  and Serial  Dilutions Performed on the Same Field Sample
      (QC Codes FLD, LDO, LD2, LSF,  LDF)
30|C|7440-22-4| | | | |T|5000| I I I I
31|W|I| 1328.0011.9540) | 1.96101
30|C|7429-90-5| | | | |T|1000| III)
31) W|I| 1308.2010.8384) | 0.8378)
30|C|7440-39-3| I | | |T|5000| Mil
31) W|I| 1493.40(1.9460) | 1.9510)
30|C|7440-41-7| | | | |T|5000| MM
31| W|I| 1313.00)0.9924) 10.9910)
1 1 1 1 |U|3.1
1.9660) | I |
1 II 1 IUI21.
0.8440) | | |
1 IMIUIll.
1.9684) | | |
MM IUI1.1
1.0010) | I I
   20|1|MAX123|F| OW) |20596IMAX12311999|09117111109||G|1.05 I 08|01568C5FD
   21|ILOWIIHS1IIS308233-01I1999I09I14)|1999|08|24|||08)30|01569D451
   22)||||||||1200)1.01||91.5|01570DE17
   90|STONES)01571E154
   92|GREY|GREY| | |MEDIUM)YES I01572EA43
   30|C|7440-22-4| |MG/KG|BDL|2.0817| I II I I I I I I I I |U|0.82||1.15671 ||(01573FD12
   31|W|C||328.00|4.2000|10.5500)1-1.2800)|||I0157409A5
   30ICI7429-90-5) IMG/KG)16227.0101)I I I I IMI I II|U|4.8| 129913.00001 I II015751DCD
   31|W|C|(308.20129992.0000)129654.00001130093.00001|||I015762CAO
   30|C|7440-39-3||MG/KG|LTC|21.9349| I I I IM MI I I |U|3.1|1105.37001||I01577400C
   31|W|C|1493.40)107.2400)1101.6400)|107.23001|||(015784DA6
   30|C|7440-41-7| |MG/KG|BDL11.0409| I I I I I I I I I I I |U|0.42| 11.4900) I II01579606A
   31|W|C||313.00|1.4900|(1.4900111.49001|||I015806CD9

   20|1|MAX123|F| BUS I I 20596 IMAX123 11999 I 09 117 1111 09 I |G| 1.05 | 08 | 01650C630
   21|ILOWj|HS1||3308233-01)1999|09|14|11999)08)24)||08|30|01651D484

                                          I|IIUI3.9I11.15671||I01655FC98
                                          I I|I01656092B
                                          I | ||U|23.1|129913.00001 I II016571C09
                                          30093.0000)|||I016582ADC
                                          I I I I|U|14.9|  1105.37001|II016593DCE
                                          2300)I||I016604B68
                                          I I |U|2.0|11.4900) I I I01579606A
                                          I|I015806CD9
      22||M II I I I |200|1.0|||91.5|01652DE4A
      30|C|7440-22-4||UG/L|BDL|10.00 I I M II I I
      31|W|C|1328.00)4.2000)10.5500)1-1.2800
      30ICI7429-90-5) IUG/L)129913.00)I I  I I  I I I
      31|W|C|1308.20129992.0000)129654.0000)
      30|C|7440-39-3||UG/L|LTC|105.37|||||||
      31|W|C|1493.40)107.2400)1101.6400)|107
      30|C|7440-41-7| |UG/L|BDL|5.00| I I I  I I  I I I
      31|W|C|1313.00)1.4900)|1.49001|1.49001
   20|1|MAX123D|F| QBE I |20596IMAX12311999|091171111111|G|1.04|08|016913BCF
   21)ILOW) I HSU|S308233-02|1999|09|14| 11999)08)24)I|08 I 30|016924A23
   22)I I I I I I I I 1200)1.0)||90.9|0169353EC
   30|C|7440-22-4||MG/KG|BDL|2.1017| I I MII I I II IIUI0.82)10.9600)| |I0169466BE
   31|W|C|1328.00)1.6400)11.6300)1-0.3900)|||1016957356
   30 1C|7429-90-5|IMG/KG)16622.74061 ||P|6.2||||IMIIU|4.8|131511.0000) ||1016968784
   31|W|C|1308.20)31993.0000)131313.0000)131227.0000)|||1016979641
   30|C|7440-39-3||MG/KG|LTC|25.1387| | |P|13.6|I I I I I I IIU|3.11 1119.6100) ||I01698AAC5
   31|W|C| 1493.40)121.4600)1118.9300)1118.4400) I I I I01699B86C
   30|C|7440-41-7||MG/KG|BDL|1.0509 I I I I I I I I II  I I |U|0.42|11.5000)I II01700CC13
   31|W|C|1313.00)1.5000)11.5000)|1.50001|||I01701D86A
          ILM05.2
                                     Appendix A-6

-------
1.2
                                                                               Appendix A
                                             Format of Records for Specific Uses (Con't)

             F| OSS\ | 20596 | MAX123 1 1999 | 09 1 17 1 11 1 14 | | G 1 1 . 01 1 08 | 01730BE3C
            . |S308233-03|1999|09|14| 1 1999 | 08 | 24 | | | 08 I 30 | 01731CC90
            20011.01 | |91.5|01732D656
           !-4| IMG/KG | |10.7212|F|10.82|P|99| | | | | |75|125| |U|0.82| |49.5400| | | (01733EBC7
            00148.84001 149.20001 |50. 58001 | | | (01734F8DC
            -5| |MG/KG|NAR|6859.9253| I I I I I I I I I I I |U|4.8| 131698.00001 | | I017350E27
            20131578.0000) 131766.00001 131750.00001 I I I I017361CF1
           '-3| IMG/KGI |326.3539|F|432.83|P|70| | | | |N|75|125| |U|3.1| 11508. 00001
            4011524.00001 11504.40001 11495. 60001 | | | (017384171
            -7| IMG/KGI |10.4290|F|10.82|P|96| MM |75|125| IUI0.42I 148.19001 | | I0173956E4
            00148.19001 |48.2000| | 4 8. 18001 | | | I0174063EB
      20|1|MAX123S
      21||LOW||HS1
      22| | | | | | |
      30|C|7440-22
      31|W|C|1328.
      30|C|7429-90
      31|W|C|1308.
      30|C|7440-39
      31|W|C||493.
      30|C|7440-41
      31|W|C||313.
      20|1|MAX123L|F| nHi| | 20596 IMAX123 11999 | 09 117 111117 |
      21)|LOW|||(S308233-04I||||1999|08|24||||(017707255
      22| | | I I I I I I I 15.01 | |91.5|017717B8D
                     |UG/L|BDL|50.00| I II II I IIII
                     1.45001 (-0.38001 (0.76001 I |
                     (UG/LI(25575.501 | (PI15II I |
                     5038.60001(5126.4000115180
                     |UG/L|LTC|111.30| I IP|6| | | |
                     22.2600|122.77001(21.75001
                     |UG/L|BDL|25.00| I I I I  I I I I I I
                                                                                017373339
                                                     (08(017696573
30)0)7440-22-4
31|W|C|1328.00
30|C|7429-90-5
31|W|C|(308.20
30|C|7440-39-3
31|W|C|(493.40
30ICI7440-41-7
I017728DDF
                                          |U|3.9| (0.61001
                                          (017739A7B
                                          I I (EIUI23.1I 15115.10001 I I I01774AE69
                                          3000| | | | (01775BCAC
                                          I I I |U|14.9| 122.26001 I I (01776DOAA
                                          I | I01777DDB9
                                          |U|2.0| 10.30001 | | (0173956E4
31|W|C|(313.00(0.19001|0.20001|0.51|||I(0174063EB

ICP-MS
1.2.1 Start of an ICP-MS Run with Record Types 10 and 16 and the First Type 20 Records

      10(1999(09117|09|06|MS|ILM05.2|ABC|TESLAB||||68-D2-0039|P2|(TEST LABSINC.(2(000001879
      1611999|09 117112|03|Y|Y|YIN I 000012114
20|1|MDL|1|MDQ| | | 1 1999
21| | I INP1I | I | |  | I | I (I I
30|C|7440-22-4| (UG/LI
30|C|7429-90-5| IUG/LI
30|C|7440-39-3| IUG/LI
30|C|7440-41-7| IUG/LI
                            | 07 1 15 | | | I
                            (00005DD31
                            MINIMI
                            II I I Ml II
                            III III III
                            IIIIMIII
      20|1|MDL|1|MDQ| I I |1999 I 07115 I  I I
      211| I |HW2| I |1999|07|15|I I I I  I(09
      30|C|7440-22-4||UG/L|I I I I I I  I I  I I
      30ICI7429-90-5I |UG/L| I I I I I I  I II I
      30|C|7440-39-3| |UG/L|I IM I II I  I I
      30|C|7440-41-7| |UG/L|I I II I I  I I  II
20|1|LRV|
30|C|7440
32 | |
34) |
30|C
32||
34| |
30|C
32||
34| |
30|C
32 | |
34| |
                                I I04I000044B9D

                                I| IUI0.40)I I I|1000065996
                                |||U|12.8|||||I0000767D1
                                I||U|3.0|||||I0000875CB
                                |||U|0.44|||||I0000983C5

                                I|04|000044B9D
                                OOI00005DD31
                                I | |U|0.41| |II I 1000065996
                                I|IUI13.8)1)11I0000767D1
                                I | |U|4.0| Mil I0000875CB
                                I||U|0.43|||||I0000983C5
      |107
      7429
      || | |
      |27.
      7440
      | | | |
      |137
      7440
      | || |
      1111
               1|LRQ| | I 11999 I 07)15 I I I   I I 04 | 0002356C2
               -22-4| I M I I I I  M M I I I I I   I M II I0002463D1
                  IT | 5 . 00 | | |  | 000256CDA
               .00|5|40000| MM 1000267591
               -90-51 I I I I I II  I I II II M I  I I II I I I0002782AD
               |  | |IT|5.00| | M000288BB6
               00|200|1000000| MM I0002994FB
               -39-3) I I I I I I I  I I I II I II II II I II I00030A211
               || IITI5.00I | |  I00031AB1A
               .00|20|100000| I I I I I00032B436
               -41-71 I I I I I I I  I I I I I II I I  I I I M I I00033C149
               || | IT | 5 . 00 | I II 00034CA52
               .0012)25000) I  I I I I00035D2DA
1.2.2 ICP-MS Instrument Tune and Calibration Standards, SO and S

      20|3|TUNEA1|1|LTS| | 26791 IMCSBOO 1 1999 | 02 I 06 I 20 | 00 | | |  |5|000917DD7
                                            Appendix A-7
                                                                            ILM05.2

-------
Appendix A
Format of Records for Specific Uses (Con't)
   21| | | | | ITUNE1I I I I 1 I I ITESLABI I I I000917DD8
   22)||||||||||1.0|||I000917DD9
   30ICI7440-41-7I | | | |T|100| I I I I I I I I I I
   31|M|I|19.0111000001110000011100000
   30|C|7439-95-4| ||||T|100|I I I I I I I I I
   31|M|I||23.99|79000||79000||790001
   31|M|I|124.991100001|10000||100001
   31|M|I|125.981110001|110001|110001
   30|C|7440-48-4| | | | |T|100| I II I I I | I |
   31|M|I||58.93|100000||100000|1100000
   30|C|7440-74-6|I I I |T|100|I I I II I I I I
   31|M|I||112.90|4000||40001|4000|||
   31|M|I|1114.90(960001|960001|96000)
   30|C|7439-92-l| | |||T|100|I I I I I I I I I I
   31|M|I||205.97|24000||24000||24000|
   31|M|11 |206.98|22000|1220001 I 220001
   31|M|I|1207.981520001|52000||52000|
                                      I I I I000917DEO
                                      II000917DE1
                                      I||I000914DE2
                                      000917DE3
                                      000917DE4
                                      000917DE5
                                      I I I I000917DE6
                                      I II000917DE7
                                      I | |I000917DE8
                                    000917DE9
                                      I000917DFO
                                      | | |I000917DF1
                                      I000917DF2
                                      I000917DF3
                                      I000917DF4
20|3|TUNEA2|1|LTS| | 26791 (MCSBOO | 1999 | 02 | 06 | 20 | 10 |
211 1 1 1 1 |TUNE2| 1 1 1  1 1 1 ITESLABI 1 1 1000917008
22) I I II I I I I I 11-01  I I I000917DD9
30 |C| 7440-41-7 | I I  I |T| 100 | | |  I I I I I I I I I
31|M|I| |9.01|100000| 1 100000 I 1 1000001
30|C|7439-95-4| | |  | |T|100| I I  I I I I I I I I I
31|M|I| |23.99|79000| |79000|  |79000| |
31|M|I| 124.99110000) 1100001  |10000| I
31 1 M| 1 1 125.98111000) 1 110001  1 11000 I I
30|C|7440-48-4| | |  | |T|100| I I  I I I I I I I I
31|M|I| |58.93|100000| (1000001 (100000
30ICI7440-74-6I I I  I I T 1 100 | | |  I I I I I I I I
31|M|I| (112.90(40001 (40001 (4000) | | |
31|M|I| (114.90(960001 (960001 (960001
30|C|7439-92-l| | |  | |T|100| I I  I I I I I I I I
31|M|I| 1205.97)240001 (240001 124000)
31 1 M| 1 1 1206.98(22000) | 22000) 1 22000)
31|M|I| 1207.98)52000) 152000) 152000)
                                                      |5|000917DD7
                                         1000001||I000917DEO
                                         II000917DE1
                                         I||(000914DE2
                                         000917DE3
                                         000917DE4
                                         000917DE5
                                         100000)||I000917DE6
                                         I | I000917DE7
                                         I | | I000917DE8
                                       000917DE9
                                         I000917DFO
                                         I|II000917DF1
                                         (000917DF2
                                         (000917DF3
                                         (000917DF4
20|3|TUNEA3111LTS| |26791(MCSBOO11999|02|06|20
2ii11111TUNES1111111ITESLABI11iooo9i7DD8
22|I I I I I I I I I|1.0| I I(000917DD9
30|C|7440-41-7| | | | |T|100| I I I I I I | | | I I
31|M|I|19.01(1000001(1000001|100000|
30|C|7439-95-4| | | ||T|100| I I I I I I I M
31|M|I|123.99)79000)179000)179000)
31|M|I||24.99|10000||10000)(100001
31|M|I|125.98)110001(110001(11000)
30|C|7440-48-4|I I I IT)100 I|I I I I I|||
311M|11 (58.93(1000001(1000001(100000
30|C|7440-74-6|| | | |T|100|| II II I I I
31|M|I||112.90|4000|(4000)(4000)|
31|M|I|(114.90(960001(960001(96000
30|C|7439-92-l|| |I|T|100|I I I I I I I I I
31|M|I||205.97|24000||240001124000
31|M|I|(206.98(220001|22000|[22000
31|M|I|1207.9815200011520001(52000
                                             | 20 |
                                                      |5|000917DD7
                                         |||(000917DEO
                                         I(000917DE1
                                         I||I000914DE2
                                         000917DE3
                                         000917DE4
                                         000917DE5
                                          | | I000917DE6
                                          I I000917DE7
                                          I | (000917DE8
                                       000917DE9
                                          000917DFO
                                          | | (000917DF1
                                          000917DF2
                                          000917DF3
                                          000917DF4
20|3|TUNEA4|1|LTS| | 26791 IMCSBOO 1 1999 I 02 | 06 | 20 | 30 | | | |
21)111(1 TUNE 4 | | | | | | | | TESLAB I M I 000917DD8
22| | I I I I I I I I 11-01 | | I000917DD9
30ICI7440-41-7) | | | |T|100| I I I I I I I I I I I I I I I I I000917DEO
31 1 M| 1 1 |9.01|100000| (1000001 (1000001 I I I I000917DE1
30|C|7439-95-4| I I I I T 1 100 | | I I I I I I I I I I I I I I I I000914DE2
                                                       5|000917DD7
       ILM05.2
                                  Appendix  A-8

-------
                                                                                     Appendix A
                                                   Format of Records for Specific Uses (Con't)
      31|M|I||23.99|79000||79000||79000|
      31|M|I| 124.99110000] 110000| I 100001
      31|M|I|125.98|110001|11000||110001
      30 1C|7440-48-41 I I I |T|100| | | | | I I I ||
      31|M|I|158.9311000001|100000|1100000
      30|C|7440-74-6 I I I I IT1100| | | I I I I I I I I
      31|M|I|1112.90140001|40001|4000||||
      31|M|I|1114.901960001|96000||960001
      30|C|7439-92-l| || | |T|100| I I I I I I I I I I
      31|M|I| 1205.971240001|240001 I 240001
      31|M|I|1206.98(220001|220001|220001
      31|M|I|1207.9815200011520001(520001
                                      000917DE3
                                      000917DE4
                                      000917DE5
                                      I I I I000917DE6
                                      I I I000917DE7
                                      I I I I000917DE8
                                    000917DE9
                                      I000917DFO
                                      I I | I000917DF1
                                      I000917DF2
                                      I000917DF3
                                      I000917DF4
20131TUNEA5|1|LTS||26791IMCSBOO|1999|02|06|20|40
21||||||TUNESi||||||ITESLABI||I000917DD8
22| I  I I I I I I I I 11.01 I I(000917DD9
30|C|7440-41-7| || | |T|100|P|0.0| | | | |  |
31|M|I|(9.01(1000001(1000001(1000001
30|C|7439-95-4| | | | |T1100|P|0.0| | | |
31|M|I||23.99|79000|1790001(79000)
31|M|I||24.99|10000||10000||10000|
31|M|I|(25.98(110001(110001|110001
30|CI 7440-48-4| || | IT1100|P|0.0| | I I
31|M|I|(58.931100000)1100000)1100000
30 |C| 74-40-74-6) | | | |T| 100 |P| 0. 0 | I I I
31|M|I| (112.90(40001 140001 14000 I I |
31|M|I|(114.90(960001(960001(96000
30 1C|7439-92-11 |1 | |T1100|P|0.0| | I I
31|M|I|(205.97(240001|24000|(24000
31|M|I|1206.98)22000)(22000)(22000
31|M|I|(207.98(520001(520001(52000
                                                       |I|5|000917DD7
                                            I I I I I I I I000917DEO
                                            II000917DE1
                                            | | | | | || (000914DE2
                                            000917DE3
                                            000917DE4
                                            000917DE5
                                            I I I I I I I I000917DE6
                                            I | I000917DE7
                                            |||||||I000917DE8
                                          000917DE9
                                            (000917DFO
                                            I I I I | I I I000917DF1
                                            I000917DF2
                                            I000917DF3
                                            I000917DF4
                                             09(061 | | (04I00128D199
                                             I I00129DD31
                                             40) | | | | I00131F8F5
                                             I |  (001320305
20(11811II STB||20596|MAX12311999|09117
211 | I INP1I |STDB11999|09 117|  | I | ITESLABI
22|| I I  I I I  I I I  U-OI I I I00130E598
30|C|7440-22-4|||I|T|0.0|||||||||||U|0
31|M|I|(107.00(0.03041|0.0374||0.0400|
30|C|7429-90-5| I  I I |T|0.0|I I  I I I I I  I I  |U|12.8| MM 1001331697
311M|11 |27.00|0.0104| 10.01361 I 0.01201 | | | (001342137
30|C|7440-39-3| I  I I IT|0.0| | |  | I I I I  I I  |U|3.0| MM (00135348D
31|M|I| 1137.001-0.0002) 10.0002) (0.00001 I I | I001363EA4
30|C|7440-41-7|||||T|0.0||(IIIIIIIIU|0.44|||||(0013751FA
31|M|11 (111.0010.0006110.00021 10.0004) | | | (001385C04
     _|1|STD||20596|MAX123|1999|09|17|
     (NP1I |STD1|1999|09)17| |I  I (TESLABI
        I I I I I (1.01 | | I00208453C
20(11
211 I  I
22|
30|C|7440-22-4| || | |T|5000| (III I I I I|U|0
31|M|I|(107.00(1.95401(1.96101 1.96601|
30|C|7429-90-5 I 1 I I |T|10001 (III I I I I|U|
311M|11 (27.00(0.83841(0.8378) 10.8440) | |
30|C|7440-39-3| |1 | |T|5000| I I  I I I I I I(U|
31|M|I|1136.0011.94601(1.95101 1.96841||1
30|C|7440-41-7| | | | |T|5000| MM I I I I(U|
31|M|
                                            09|11| | | (04I00206314E
                                            I | (002073CD5
             |111.00|0.9924|(0.9910)
1-00101 I I I I
                                        401 | | | | (002139157
                                        | (002149B6E
                                      12.81 | I I | I00215ADE2
                                        | I00216B7EC
                                      3.0| | I | | I00219E77D
                                          (00220F18F
                                      0.441 | | I I 1002210410
                                           002220E25
1.2.3 Field Samples
                        | I 20596 IMAX123 11999 | 09 117 | 09 | 06 I I ML 1100 | 04 | 00128D199
      21||||HW2||S308233-01|1999|09|17||1999|09|16|TESLAB||09|30|00129DD31
      22| | | | | I I  11.25) 150)1.0) I IO.OI00130E598
      30ICI7440-22-4) |UG/L|LTC|0.6625| I I I I  I I I I I I I |U|0.41| 10.53001  I| (00131F8F5
                                            Appendix A-9
                                                                            ILM05.2

-------
   Appendix A
   Format of Records for Specific Uses
                                  (Con't)
                                         (001320305
                                         IUI13.8)145.1000)I I (001331697
                                         I I 1001342137
                                       I  IUI4.0)(8.80001||I00135348D
                                         (001363EA4
                                         I|U|0.43|(0.32101||I0013751FA
                                         I001385C04

                   I 20596IMAX12311999 I 09 117|09|06||ML|20|04|00128D199
21)| | INP1I|S308234-01|1999|09|17|11999)09 116|TESLAB||09|30|00129DD31
22|||||||||(20)1.0)|(O.OI00130E598
      31|M|I||107.00|0.5300||0.5300110.53001
      30|C|7429-90-5||UG/L|156.37501 I I I I I I I I
      31|M|I||27.00|45.1000||45.10001|45.10001
      30|C|7440-39-3||UG/L| 111.00001 I I I I I I I I
      31|M|I||137.00|8.8000|18.8000)|8. 80001
      30|C|7440-41-7| |UG/LIBDL|1.000| I I M I I I
      31|M|I||111.00|0.3210||0.3210|(0.32101
      30|C|7440-22-4|(UG/DLTC | 0. 5300 |Mill)  I
      31|M|I|(107.00(0.5300)fo.5300)10.5300)
      30|C|7429-90-5|IUG/L)(45.1000111(11(11
      31|M|I|(27.00(45.1000)(45.1000)145.1000)
      30|C|7440-39-3||UG/LILTC|8.8000|I I I I I I
      31|M|I|(137.00(8.8000118.8000)18.8000)
      30|C|7440-41-7| |UG/L|BDL|1.000| I I I I I I I
      31|M|I|1111.0010.3210110.3210110.32101
I | |U|0.40| (0.53001
(001320305
|U|12.8| (45.10001 |
I | (001342137
I I |U|3.0| (8.8000) |
I001363EA4
I IUI0.44I 10.3210) |
I001385C04
                                                            | (00131F8F5

                                                            (001331697

                                                            I00135348D

                                                            I0013751FA
2.0   MERCURY
2.1
Start of a Mercury Run for Water Samples with Record Types 10 and 16 and the First Type 20
Records

10|1999|09|09|08|44|CV|ILM05.2|ABC|TESLAB||||68-D2-0039|M3|(TEST LABS INC.|6(0000018F7
16|1999|09|09|14(34|N|||1000012099
      20|1|MDL|1|MDQ||||1999|07|15|
      21)|||CW1||11999107(15(1 I III I
      30ICI7439-97-6)|UG/L|I I  I I II  II
                              I | ) UI000044AEB
                              (000053CD5
                              I I I I(UI0.042)|||
     I (0000658F4
20|1|LRV|1|LRQ| I I(1999)09|09|
30|C|7439-97-6|I I I I I I I I I I I I I  I
32)|||||||||||||(000087D02
34|||(253.70(0.2(51 I I II I00009852D
                                       |1|0000666A6
                                       || I II0000773CB
2.1.1 Start of a Mercury Run for Soil Samples with Record Types 10 and 16 and the First Type 20
      Records

      10|1999|09|09|08|44|CV|ILM05.2|ABC|TESLAB||||68-D2-0039|M3|(TEST LABS INC.|6|0000018F7
      1611999 I 09|09 114|34|N|I I 1000012099

      20|1|MDL|F|MDQ| || |1999|07|16|I I I I |1|000074AEB
      21| || |CS1| | |1999|07|16|((III I 09|00|000083CD5
      30|C|7439-97-6|IMG/KG I I I I  I I  I I I I I I I I |U|0.00921 I  II I(0000958F4

      20|1|LRV|1|LRQ| I I (1999(09(091 I I I I UI0001066A6
      30|C|7439-97-6|I I I I I  I I I I I  I I  I I I I I I I I I  I (0001173CB
      32| || I I| I| | | | |  | |I000127D02
      34) ||(253.70(0.2(51 I  I I II00013852D

2.2   Mercury Instrument Calibration Standards: Blank (SO) and Four Other Standards

      20|l|aS|l|STB||20596IMAX12311999 I 09 I 09 I 08 I 44||IUI00010936F
      21| || |CS1||OPPB11999 I 09 I 09|  ||| |TESLAB| |07|00 I000119FOC
      22| || || | |  | | | (1.01 | |I00012A773
      30|C|7439-97-6| ||| |T|0.0|  I I  I I I I I I I |U|0.018|10.01221 I|I00013BAD9
      31|W|C| 1253.70)0.0122)I I I  I I  I II00014C4EC

      20|1|S0.2|1|STC||20596IMAX123|1999|09|09|08|48|||(1(000150392
      21| |||CS1| |0.2PPB|1999|09|09|I I I(TESLAB)I 07|00|00016DF8F
      22)|I)||||||11.0)||I00017E7F6
          ILM05.2
                                  Appendix A-10

-------
                                                                                     Appendix A
                                                    Format  of  Records  for  Specific Uses  (Con't)
      30|C|7439-97-6| I I | |T|0.2| I I I I I I I I I |U|0.018| |0. 08961 | | I00018FB5E
      31|W|C| |253.70|0.0896| I I  I I I I I 1000190571
                       | 20596 | MAX123 1 1999 I 09 | 09 | 08 | 53 | | | |1|000201412
      21| | | |CS1| |1.0PPB|1999|09|09| I I | |TESLAB| | 07 | 00 | 00021200E
      22| | | | | | | | | | |1.0| | | 1000222875
      30|C|7439-97-6| |  | | |T|1.0| I I I  I I I I I I  |U|0.018| |1. 01281 | | I000233BDC
      31|W|C| 1253.7011.01281 I I  I I I I  I I0002445EF
                       |20596|MAX123|1999|09|09|08|57| |  | |1|000255495
      21| | | |CS1| | 2. OPPB 1 1999 | 09 | 09 | I I I |TESLAB| | 07 | 00 | 000266092
      22| | | | | | |  | | |  |1.0| | | I0002768F9
      30|C|7439-97-6| I  I I |T|2.0| I I I I I I I I I |U|0.018| |2. 00551 | | I000287C61
      31IWICI 1253.7012.0055) I I  I I I I I 1000298674

      20]1|^E|1|STD| | 20596 | MAX123 1 1999 | 09 | 09 | 09 I 01 1 |  | |1|000309513
      21| | | |CS1| | 5. OPPB | 1999 | 09 | 09 | | | | |TESLAB| | 07 | 00 | 00031A113
      22| | | | | | |  | | |  |1.0| | | I00032A97A
      30 |C| 7439-97-6 | |  | | | T | 5 . 0 I P | 0. 9997 | | | |  | |  | | |U| 0 . 018 | | 4. 99521 | | I00033BCE5
      31|W|C| | 253. 70 | 4. 9952 | I I  I I! I I I00034C6F8

2.3   Spike Sample  Recovery and Duplicates Performed on Different Samples
      (QC Codes FLD, LSF,  FLD,  LD2)
      20|1|MAX123|F| OM| | 20596 IMAX123 1 1999 | 09 | 09 1 13 I 20 | I G| 0 .20 1 1 1 002106798
      21| | LOW | |CS1| |S308233-01|1999|09|09| |1999|08|24| | | 07 | 00 I 0021175EF
      22| | | | | | | | | 110011.01 | I91.5I002127FB4
      30|C|7439-97-6| |MG/KG|BDL| 0 . 1093 | I I  I I I I I I I I I |U|0. 00921 |0.0049| | | I002159ECO
      31|W|C| |253.70|0.0049| I I I I I I I I00216A8E3
      20|1|MAX123S|F| DH3I | 20596 IMAX123 1 1999 | 09 | 09 1 13 I 25 I I G| 0.20 1 1 1 00229534B
      21| | LOW | |CS1| |S308233-03|1999|09|09| |1999|08|24| | | 07 | 00 | 0023061A2
      22| | | | | | | | | |100|1.0| | |91.5|002316B67
      30 |C| 7439-97-6 | IMG/KG | | 0. 5664 I F| 0 . 55 | P| 103 | MM | 75 1 125 I I U| 0.00921 1 1.03661 | | I00232807A
      31|W|C| |253.70|1.0366| II II | |  | I002338A9D
      20|1|MAX126|F| 9*31 | 20596 | MAX123 1 1999 | 09 | 09 1 13 | 30 | | G| 0.20 1 1 1 00217B9F5
      21| | LOW | |CS1| |S308233-06|1999|09|09| 1 1999 | 08 I 24 | | | 07 | 00 | 00218C84C
      22 M M M I I I  |100|1.0| | |85.6|00219D211
      30|C|7439-97-6| |MG/KG| 11.50531 M M M M M I |U|0. 00921  |2.5771| | | I00222F11D
      31|W|C| |253.70|2.5771| M M M I I00223FB40
      20|1|MAX126D|F| BOB | | 20596 IMAX123 1 1999 | 09 | 09 1 13 | 35 I I G| 0.20 1 1 1 002240C9D
      21| | LOW |  |CS1| |S308233-07|1999|09|09| |1999|08|24M I 07 | 00 | 002251AF4
      22M M M  M I  110011.01 | | 85.1 1 0022624BC
      30 |C| 7439-97-61 IMG/KG |BDL| 0 . 1175 |  | |P|200| | |L| 0.0383 I |  | |*|U| 0.0092| 10.0028 I | | 1002273795
      31|W|C| |253.70|0.0028| M M M I  I0022841B9

2.4   Duplicates and Spike Sample Recovery Performed on the Same Sample
      (QC Codes FLD, LD2, LSF)

      20|1|MAX126|F| fffl!| | 20596 IMAX123 1 1999 | 09 | 09 1 16 1 10 | | G| 0 .20 1 1 1 002106798
      211 ILOWI  |CS1| |S308233-06|1999|09|09| |1999|08|24| | | 07 | 00 I 0021175EF
      22 M M M  M M 100 1 1 . 0 M I 91 . 5 | 002127FB4
      30|C|7439-97-6| |MG/KG| 10.64291 M I  M M M M |U|0. 00921 |1. 17651 |  | I002159ECO
      31|W|C| |  253. 70 | 1.1765 | M M M I  I00216A8E3

      20|1|MAX126D|F| GEB I | 20596 IMAX123 1 1999 | 09 | 09 1 16 1 15 | |G| 0 . 20 1 1 1 002240C9D
      211 ILOW)  |CS1| |S308233-07|1999|09|09| |1999|08|24| | | 07 | 00 | 002251AF4
      22|||MIIIII100|1.0||| 90. 9 | 0022624BC
      30 |C | 7439-97-6 | IMG/KG | | 0.23421 | | P | 94 ||  | L | 0 . 0364 I | | |*|U| 0.00921 | 0.42861 | |  1002273795
      31|W|C| |  253. 70 | 0.4286 | I  M I I I I  I0022841B9


                                           Appendix A- 11                          ILM05.2

-------
   Appendix A
   Format of Records  for  Specific Uses  (Con't)

      20|1|MAX126S|F| BHS| |20596IMAX12311999|09|09116|20 I IGI 0.2011100229534B
      21||LOW||CS1||S308233-08|1999|09|09||1999|08|24|||07|00|0023061A2
      22|||||||||110011.01|I91.5I002316B67
      30 1C I 7439-97-6||MG/KG||0.9710|F|0.55|PI 60|||||N|751125 I|U|0.0092111.7769|||I00232807A
      31|W|C|1253.7011.77691 I I I I I I II002338A9D

2.5   Initial  Calibration Verification  (ICV) with LVM QC Code

      20|l|ICVlA|l|iIBB51| | 20596 IMAX123 11999 | 09 | 09 | 09 | 06 | | | |1|00035D687
      21||||CS1||ICV-5|1999|09|09|||07|00|ICF(0791)||07|00|00036E25E
      22||I||||||||2.0|||I00037EAC6
      30|C|7439-97-6||UG/L|| 4.91|T|4.9|P|100 |I I I I |80.01120.01|U|0.018|12.45591 I I I00038FFDO
      31|W|C||253.70|2.4559|I I I I I I II0003909FC

2.6   Laboratory Control Sample (Solid) with LCM QC Code

      20|1|LCSSC3|F| OSS\ I 20596IMAX12311999 I 09|09112 I 24| |G|0.20111001256DBA
      21||||CS1||LCSHG|1999|09|09|||||QAL-0287||07|00|001267B1B
      22||||I||||110011.01||1001278443    ,
      30|C|7439-97-61|MG/KG||4.6|T|4.2|P|110||||||2.8|6.0||U|0.0092||9.20001||I00128996D
      31|W|C| |253.70|2.7719| I I I I I I I I00129A39A

3.0   CYANIDE

3.1   Start of a Cyanide Run with Record Types 10 and 16 and the First Type 20 Records

      10|1999|09|01|14|09|AS|ILM05.2|ABC|TESLAB| | | |68-D2-0039|C11  I TEST LABS INC.|7|00000189C
      1611999|09 I 01115|03|Y|||1000012033

      20|1|MDL|1|MDQ|  |||1999|07|15| I I I  I|1|000044A74
      21|||INP1I||1999|07|15|||||||10|30|000053CD5
      30ICI57-12-5IIUG/LII I I I I I I I I I I I I  IU11.7||||||0000656DC

      20|1|MDL|1|MDQ|| ||1999|07|15|I I I  I|1|000044A74
      211|| |OW11 |11999|07115||||||110|30 I000053CD5
      30|C|57-12-5|IUG/LII I I I I I I I I I I I I  |U|1.8| I  I I I I0000656DC

      20|1|MDL|F|MDQ|I||1999|07|16|MM|1|000044A74
      21||| |DS2| | |1999|07|16|IM MI 07|45|000053CD5
      30ICI57-12-5I (MG/KGI I I I I I M I I I I I  I |U|0.092| MM  I0000656DC

      20|1|LRV|1|LRQ|  I I|1999|09|01| MM 11)000066486
      30|C|57-12-5| M  M M M M M M M M M M I000076FDA
      32M I I I I I I I |IT|45.00| | | 1000087917
      34| | ! 1620.00110(4001 MM 1000098169

3.2   Cyanide Instrument  Calibration Standards: Blank (SO) and Five Other Standards

      20|1|SH|1|STB||20596IMAX12311999|09|01114|09|I I 111000108FA1
      21 M I INPII | OPPBI M M 11 ITESLABI 11  |OOOH9B3E
      22M M M M M ll.OM I I00012A3A5
      30|C|57-12-5|||||T|0.0|||||||||||U|1.7||0.35431||I00013B48B
      31|W|C||620.00|0.3543|M M MII00014BD34

      20|l|Smil|STC||20596 IMAX12311999|09|01114110MI UI00015CB95
      21M I INPII I10PPBI I I I M I ITESLABI I  I I00016D763
      22|  M I I I II M 11-01||I00017DFCA
      30|C|57-12-5| | | | ITI10.0I I I I M I I I  I |U|1.7| 111.17001 | | I00018FOD2
      31|W|C| |620.00|11.1700|IMI I I I(00019F97B

      20|1|SEH|1|STD||20596 IMAX12311999|09|011141111| | |1|0002007EO
      21M iINPII14OPPBI111111ITESLABI111000211331
      22||M|| M I I|1.0|I II000221C18

          ILM05.2                       Appendix A-12

-------
                                                                                     Appendix A
                                                    Format of Records  for  Specific Uses  (Con't)
      30 |C| 57-12-5 | ||||T|40.0|||||||||||U|1.7| | 38. 40001 | | I000232D23
      31|W|C| |620.00|38.4000| I I I I M I I0002435CC
      20| 1| am 11 1 STD | I 20596 | MAX123 1 1999 | 09 | 01 1 14 1 12 | | | UI00025445F
      211 | | INPII IIOOPPBI i i i | | | ITESLABI i i  I00026505D
      22| I I I I I I I I  I 11.01 I I I0002758C4
      30 |C| 57-12-5 I I I I IT| 100.01 | | | |  | |  | | I  IU |1.7| | 99. 7400) | | | 000232D23
      31 | W|C| |620.00|99.7400| | I | | | |  I I0002972A5

      20|1|EBBEI|1|STD| | 20596 | MAX123 1 1999 | 09 I 01 1 14 1 12 | | | |1|000308139
      211 1 1 INPII I200PPB) 1 1 1 1 1 1 ITESLABI 1 1  IOOOSISDSS
      22| | | I I II I I  I 11.01 | | I00032959F
      30|C|57-12-5| | | | |T|200.0| I I I I  I I  I I I  I U 1 1 . 7 | |201.3000| | | | 00033A6D8
      31|W|C| |620.00|201.3000| I I I I I  I I  I00034AF81

      20|1|^EE|1|STD| | 20596 IMAX123 1 1999 | 09 | 01 1 14 1 13 I I | |1|00035BE18
      211 i i INPII 1400 PPBI i i i i i i ITESLABI i  i I00036CA19
      22| I I I I I I I I  I 11.01 I I I00037D280
      30 |C | 57-12-5 | I I I |T| 400. 0| P| 1.0000 I  ||||III|U|1.7| | 399. 50001 | | | 00038E3BB
      31|W|C| |620.00|399.5000| I I I I I  I I  I00039EC64

3.3   Preparation Blank  (Soil)  with LRB QC Code

      20|1|PBSD1|F|BSB| I 20596 (MAX123 1 1999 | 09 | 01 1 14 | 23 | |G| 1 . 00 1 1 1 000928FAO
      21| | | |DS2| | PB 1 1999 | 08 | 30 | Mill I 08 | 30 | 000939A40
      221 I | | | | | | |  15011.01 | | I00094A30C
      30|C|57-12-5| |MG/KG|BDL|1.0000|  I I I  I I I  I I I I I  |U|0.092| | -0.00301 I I I 00095B433
      31|W|C| | 620. 00 | -0.0030 | I I I I I I  I I00096BE6F

3.4   Laboratory Control Sample (Soil) with LCM QC Code

      20| HLCSSD1 |F| BSBI | 20596 IMAX123 | 1999|09|01|14|24| |G| 1.00| H00097CF4D
      21| | | |DS2| |LCSCN|1999|08|30| |  I I  IQAL-06891 | 08 I 30 | 00098DCBO
      22| | | | | | | | |  150)1.0) | | I00099E57C
      30 | C| 57-12-5 | IMG/KG | |5.0|T|5.6|P|89||||||4.3|6.9||O|0.092| 1 100. 09331 | | I00100F89B
      31|W|C| |620.00|100.0933| I I I I I  I I  1001010315

3.5   Continuing Calibration Verification (CCV) with LVC  QC Code

      20|l|CCVll|l|flHS| | 20596 | MAX123 1 1999 | 09 | 01 1 14 | 30 | | | |1|0015045A3
      211 1 1 INPII 1200 PPBI i M 1 1 1 ITESLABI 1  1 iooisi5iA2
      22| I I I I I I I I  I 11.01 I I I001525A09
      30 |C| 57-12-5 | IUG/LI 1 188 . 48 | T | 200 . 0 I P I 94 | | I  I I | 85. 0 1 115.01 |U| 1.7 | 1 188. 47721 | | I001536E87
      31|W|C| 1620.00)188.47721 I I I I I  I I  1001547916

3.6   Spike Sample Recovery and Post Distillation Spike  Sample Recovery Performed on the Same
      Sample (QC Codes FLD,  PDO,  LSF,  PDF)
      20|1|MAX123|F| OTWI | 20596 IMAX123 I 1999 | 09 I 01 1 14 | 35 | |G| 1 . 06 | 1 1 001955D8E
      21| |LOW| |DS2| |S308233-01|1999|08|30| |1999|08|24| | | 08 | 30 | 001966BDF
      22| | |  | |  | | | | I50|1.0| | |71.0|001977578
      30|C|57-12-5| |MG/KG|LTC|0.2952| I I I I I I I I I  I I |U|0.092| |4. 44411 | | 1002009309
      31 |W|C |  | 620.00 | 4. 4441| | | | | I I I I002019D4B

      20|1|MAX123|F| S3S| | 20596 IMAX123 1 1999 I 09 I 01 1 14 | 35 | | G| 1 . 06 1 1 1 00202AE62
      21| | LOW | |DS2| |S308233-01|1999|08|30| |1999|08|24| | | 08 | 30 | 00203BCB3
      22| | |  | I  I I I I I50|1.0| | |71.0|00204C64C
      30ICI57-12-5I |UG/L|LTC|4.44| I I  I I I I I I I I I IUI1.4I |4. 44411 | | | 00207E3DD
      31|W|C|  |620. 00|4.4441| I I I I I I I I 00208EE1F
      20|1|MAX123S|F| 0331 | 20596 IMAX123 | 1999 I 09 I 01 1 14 | 36 | | G| 1 .02 1 1 1 00209FF7A
      21| | LOW |  |DS2| |S308233-02|1999|08|30| |1999|08|24| | | 08 | 30 | 002100DCB
      22) | |  | | |  | | | I50|1.0| | |71.0|002111767

                                           Appendix A-13                          ILM05.2

-------
Appendix A
Format of Records for Specific Uses (Con't)

   30|C|57-12-51 IMG/KG||4.6341|F|6.90|P|63|| || |N|75|125| |U|0.0921167.1210 I I||0021228D6
   31|W|C||620.00|67.1210|I I I I I I I 1002133324

   20|1|MAX123A|F| 13331 | 20596IMAX12311999 I 09|01114|37||G|1.061110021444AD
   21||LOW|| ||S308233-03|1999|08|30||1999|08|24| I I II0021552FE
   22| | | | ||| || 15011.01 I|71.0|002165C98
   30|C|57-12-5| IOG/LI|21.23|Ft 20.0 IP|84||I I I|| ||U11.4||21.2279|||I0021770CO
   31|W|C||620.00|21.2279|I I I I II I I002187B4E
       ILM05.2                       Appendix A-14

-------
APPENDIX B - Modified Analysis
          Appendix B-l                          ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                        Appendix  B-2

-------
                        Appendix B - Modified Analysis

                               Table of Contents

Section                                                                    page


1.0   SCOPE AND APPLICATION 	  5

2.0   SUMMARY OF METHOD	5

3.0   DEFINITIONS 	  5

4.0   INTERFERENCES 	  6
      4.1   Spectral Interferences   	   6
      4.2   Nonspectral Interference s  	   7

5.0   SAFETY   	7

6.0   EQUIPMENT AND SUPPLIES   	   8
      6.1   Glassware/Labware  	  8
      6.2   Atomic Absorption  Spectrophotometer  	  8

7.0   REAGENTS AND STANDARDS   	   9
      7.1   Reagents   	9
      7.2   Standards	9

8.0   SAMPLE COLLECTION, PRESERVATION,  AND STORAGE    	 11
      8.1   Sample Collection  and Preservation    	 11
      8.2   Procedure  for Sample Storage   	 11
      8.3   Procedure  for Sample Digestate Storage    	 11
      8.4   Contract Required  Holding Time   	 11

9.0   CALIBRATION AND  STANDARDIZATION 	  12
      9.1   Instrument Operating Conditions  	  12
      9.2   Graphite Furnace Atomic Absorption  (GFAA) Instrument
            Calibration Procedur e	12

10.0  PROCEDURE	13
      10.1  Sample Preparation  	  13
      10.2  Sample Analysis	14

11.0  DATA ANALYSIS AND CALCULATIONS   	16
      11.1  Water/Aqueous Sample Calculation    	 16
      11.2  Soil Sample Calculatio n	16
      11.3  Corrections For Sample Dilutions    	 17

12.0  QUALITY CONTROL	17

13.0  METHOD PERFORMANCE   	  17

14.0  POLLUTION PREVENTION   	 17

15.0  WASTE MANAGEMENT  	17

16.0  REFERENCES   	17

17.0  TABLES/DIAGRAMS/FLOWCHARTS   	 17
                                  Appendix B-3                           ILM05.2

-------
                      THIS PAGE INTENTIONALLY LEFT BLANK
ILM05.2                       Appendix B-4

-------
                                                    Appendix B  — Sections 1-3
                                                           Scope and Application
MODIFIED ANALYSIS
The Contractor may be requested by USEPA to perform modified analyses.  These
modifications will be within the scope of this SOW and may include, but are
not limited to, analysis of additional analytes and/or lower quantitation
limits.  These requests will be made by the USEPA Regional CLP Project Officer
(CLP PO), USEPA OERR Analytical Operations/Data Quality Center  (AOC) Inorganic
Program Manager  (PM), and USEPA Contracting Officer  (CO) in writing, prior to
sample scheduling.  If the Contractor voluntarily elects to perform these
modified analyses, these analyses will be performed with no increase in per
sample price.  All contract requirements specified in the SOW/Specifications
will remain in effect unless the USEPA CO provides written approval for the
modification(s) and a waiver for associated defects.  The USEPA CO approval
must be obtained prior to sample scheduling.

GRAPHITE FURNACE ATOMIC ABSORPTION METHOD

1.0   SCOPE AND APPLICATION

      This method  is a graphite furnace atomic absorption spectroscopy
      procedure  that is used to analyze water, sediment, sludge,  and soil
      samples  taken from hazardous waste sites.  The  following metals:
      arsenic, lead, selenium, and thallium that are  contained  in the Target
      Analyte  List (TAL) in Exhibit C may be  quantitated by the Graphite
      Furnace  Atomic Absorption  (GFAA) method.

2.0   SUMMARY  OF METHOD

      Water  and  soil samples are treated with acids  and heat to solubilize the
      metals present.  These digestates are then analyzed for trace metals by
      the Graphite Furnace Atomic Absorption  (GFAA)  spectroscopic technique.
      In this  technique, a tube of graphite is located in the sample
      compartment  of the Atomic Absorption  (AA) spectrometer, with the  light
      passing  through it.  A small volume of  sample  solution is quantitatively
      placed into  the tube, normally through  a sample injection hole located
      in the center of the tube wall. The tube is heated through  a programmed
      temperature  sequence until finally the  analyte  present in the sample is
      dissociated  into atoms and atomic absorption occurs.

3.0   DEFINITIONS

      See Exhibit  G for a complete list of definitions.
                                  Appendix B-5                           ILM05.2

-------
Appendix B — Section 4
Interferences

4.0   INTERFERENCES

      Several types of interference effects may contribute to inaccuracies in
      the determination of trace elements in water and soil/sediments.
      Dissolved elements are determined in filtered and acidified samples.
      Appropriate steps must be taken in all analyses to ensure that potential
      interferences are taken into account.  This is especially true when
      dissolved solids exceed 1500 milligrams per Liter  (mg/L).  In addition,
      total elements are determined after appropriate digestion procedures are
      performed.  Since digestion techniques increase the dissolved solids
      content of the samples, appropriate steps must be taken to correct for
      potential interference effects.

      Interferences from the Graphite Furnace Atomic Absorption  (GFAA)
      technique can be divided into two broad categories, spectral and
      nonspectral interferences.  Spectral interferences are those resulting
      from light absorption by molecules or by atoms other than those of the
      analyte element; that is, spectral interference exists if the atomic
      absorption profile of an element overlaps the emission line of another.
      Nonspectral interferences are those which affect the production or
      availability of analyte atoms which create the measured atomic
      absorption.

4.1   Spectral Interferences

4.1.1    Emission Interference - this interference arises when the intense
         light emitted by the hot graphite tube reaches  the instrument's light
         detector, the Photomultiplier Tube  (PMT).  This problem is manifested
         by  increased signal variability  (noise) which degrades analytical
         performance.  In severe circumstances, emission interference may
         temporarily blind the PMT, resulting in erratic, meaningless readings
         at  atomization.

4.1.2    Background Absorption  - this is the most severe spectral
         interference encountered with graphite furnace  analyses.  Background
         absorption is a nonspecific attenuation of light at the analyte
         wavelength caused by matrix components in the sample.  Unlike  atomic
         absorption, background absorption is broad band, sometimes covering
         tens or even hundreds of nanometers.  This broad band absorption
         normally is due to molecular absorption or light scattering caused by
         undissociated sample matrix components in the light path at
         atomization.  Since background absorption is broad band, the chance
         of  overlap with a desired analyte wavelength is significant.

4.1.3    Emission interference is controlled by primarily by spectrometer
         optical design.  Techniques for controlling and reducing background
         absorption include matrix modification  (sample  treatment) and  optical
         background correction.  Through matrix modification, a  reagent or
         "matrix modifier" is added to the sample or standard.   The matrix
         modifier is selected to generate either an increased matrix
         volatility or decreased analyte volatility.  One type of background
         correction, Zeeman, can correct  for higher and  more  spectrally
         complicated background absorption and provide more precise and
         accurate analytical results.  Zeeman background correction uses  the
         principle that  the  electronic energy levels of  an atom  placed  in  a
         strong magnetic  field are changed thereby changing the  atomic
         spectra; the  spectral nature of  background absorption,  on  the  other
         hand is unaffected  by a magnetic  field.
 ILM05.2                        Appendix B-6

-------
                                                   Appendix B — Sections 4 & 5
                                                                         Safety
4.2   Nonspectral Interferences
      In order for atomic absorption to occur, free atoms of the analyte
      element must be present in the spectrometer light path.  Nonspectral
      interferences result when diverse components in the sample matrix
      inhibit the formation of free analyte atoms.  An often used approach to
      compensate for nonspectral interferences is known as the "Method of
      Standard Additions".

5.0   SAFETY

      See Sectioq 1.14 in Exhibit D - Introduction to Analytical Methods.
                               Appendix B-7                              ILM05.2

-------
Appendix B — Section 6
Equipment and Supplies

6.0   EQUIPMENT AND SUPPLIES

      Brand names, suppliers, and part numbers are for illustrative purposes
      only.  No endorsement is implied.  Equivalent performance may be
      achieved using equipment and supplies other than those specified here,
      however, a demonstration of equivalent performance meeting the
      requirements of this Statement of Work  (SOW) is the responsibility of
      the Contractor.  The Contractor shall document any use of alternate
      equipment or supplies in the Sample Delivery Group  (SDG) Narrative.

6.1   Glassware/Labware

6.1.1    250 milliliter  (mL) beaker or other appropriate vessel

6.1.2    Watch glasses

6.1.3    Funnels

6.1.4    Graduated cylinders

6.1.5    Various volumetric flasks  (Type A)

6.1.6    Thermometer that covers a range of 0-200°C

6.1.7    Whatman No. 42  filter paper or equivalent

6.1.8    Hot plate, block digester, or other heating source capable of
         maintaining  92-95°C

6.1.9    Balances - Analytical Balance, 300 gram (g) capacity, and minimum
         ą0.01 g.

6.2      Atomic Absorption  Spectrophotometer  - with graphite  furnace  atomizer
         and background  correction.  Hollow Cathode Lamp  (HCL) and/or
         Electrodeless Discharge Lamp  (EDL)."
 ILM05.2                       Appendix B-8

-------
                                                        Appendix B  — Section 7
                                                         Reagents and Standards

7.0   REAGENTS AND STANDARDS

7.1   Reagents

      Acids used in the preparation of standards and for sample processing
      must be ultra-high purity grade or equivalent.   (Redistilled  acids are
      acceptable.)

7.1.1    Reagent water - The purity of this water must be equivalent to ASTM
         Type II water  (ASTM D1193-77).  Use this preparation  for all
         reagents, standards, and dilutions of solutions.

7.1.2    Nitric acid - Concentrated  (specific gravity 1.41).

7.1.3    Nitric acid, 5%  (v/v) - Add 50 milliliters  (mL) cone. HNO  3 to 500 mL
         reagent water; dilute to 1 Liter  (L).

7.1.4    Hydrochloric acid - Concentrated  (specific gravity  1.19).

7.1.5    Hydrogen peroxide  (30%)

7.1.6    Matrix Modifiers

7.1.6.1     Ammonium Phosphate solution  (40%): Dissolve 40 grams  (g) of
            ammonium phosphate,  (NH4)2PO4  (analytical reagent grade) in reagent
            water and dilute  to  100 mL.

7.1.6.2     Calcium Nitrate solution: Dissolve 11.8 g of calcium nitrate,
            Ca(N03)2 =ť 4H2O  (analytical reagent grade) in reagent water and
            dilute to 100 mL.   1 mL -  20 mg Ca.

7.1.6.3     Lanthanum Nitrate solution: Dissolve 58.64 g of  American Chemical
            Society  (ACS) reagent grade 2.  La2 O3  in  100 mL  cone. HNO3 and
            dilute to 1000 mL with reagent water.   1 mL =  50 mg La.

7.1.6.4     Nickel Nitrate solution, 5%: Dissolve  24.780 g of  ACS reagent
            grade Ni(NO3)2 '*'• 6H20 in reagent water  and make up  to 100 mL.

7.1.6.5     Nickel Nitrate solution, 1%: Dilute 20 mL of the 5% nickel nitrate
            to  100 mL with  reagent water.

7.2   Standards

7.2.1    Introduction

         The Contractor must provide all standards  to be used  with  this
         contract.  These standards may be used only after they have been
         certified according  to  the procedure  in Exhibit E,  Section 8.0.  The
         Contractor must  be able to verify that the  standards  are certified.
         Manufacturer's certificates of analysis must be retained by the
         Contractor and presented upon request.

7.2.2    Stock  Standard Solutions

7.2.2.1     Stock standard solutions may be purchased or prepared from ultra
            high purity grade chemicals or metals.  All salts  must  be dried
            for 1 hour  at 105"C  unless otherwise specified.

             (CAUTION; Many  metal salts are extremely toxic and may  be  fatal if
            swallowed.  Wash  hands thoroughly  after handling.)  Typical  stock
            solution preparation procedures  follow.
                               Appendix B-9                              ILM05.2

-------
Appendix B — Section 7
Reagents and Standards  (Con't)

7.2.2.2     Arsenic solution, stock  [1 mL = 1 mg As  (1000 mg/1)]  -  Dissolve
            1.320 g of  As2O3 in 100 mL of reagent water containing  0.4 g NaOH.
            Acidify the solution with 20 mL cone.  HNO  3 and dilute  to 1 L.

7.2.2.3     Lead solution,  stock [1 mL = 1 mg Pb  (1000 mg/L)]  - Dissolve  1.599
            g of Pb(NO3)2 in reagent water.  When solution is complete,
            acidify with 10 mL  of  cone. HNO 3  and dilute to 1 L with reagent
            water.

7.2.2.4     Selenium solution,  stock  [1 mL = 1 mg  Se  (1000 mg/L)] - Dissolve
            0.3453 g of H2Se03  (actual assay 94.6%) in reagent  water and make
            up  to 2.00 mL.

7.2.2.5     Thallium solution stock  [1 mL - 1 mg Tl  (1000 mg/L)]  -  Dissolve
            1.303 g of  T1N03  in reagent water.  Add  10 mL  of cone,  nitric acid
            and dilute  to 1 L with reagent water.

7.2.3    Working Standards

7.2.3.1     Secondary Dilution  Standards

            Prepare dilutions of the stock solution  to be  used as calibration
            standards at the  time  of analysis.   These  solutions are also to be
            used  for "standard  additions".  The  calibration  standards must be
            prepared using the  same type  of acid and at the  same concentration
            as  will result in the  sample  to be  analyzed after sample
            preparation.

7.2.3.2     Calibration Blank

            Prepared by diluting  1 mL of  (1+1)  HNO 3 and 2 mL  30% H202  to  100 mL
            with  reagent water.
 ILM05.2                       Appendix B-10

-------
                                                        Appendix B — Section 8
                                   Sample Collection,  Preservation,  and Storage

8.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE

8.1   Sample Collection and Preservation

      All samples must be collected in glass or polyethylene containers.
      Water/aqueous samples must be preserved with nitric acid to pH less than
      2 immediately after collection.  All samples must be iced or
      refrigerated at 4"C (ą2"C) from the time of collection until digestion.

8.1.1    Dissolved Metals

         For the"determination of dissolved metals, the sample must be
         filtered through a 0.45 micrometer  (pm) pore diameter membrane filter
         at the time of collection or as soon as possible.  Use a portion of
         the sample to rinse the filter  flask, discard this portion, and
         collect the required volume of  filtrate.  Preserve the filtrate with
         nitric acid to pH less than 2 immediately after filtration.

8.2   Procedure for Sample Storage

      The samples must be protected from light and refrigerated at 4"C  (ą2°C)
      from the time of receipt until 60  days after delivery of a complete,
      reconciled data package to USEPA.  After 60 days the samples may be
      disposed of in a manner that complies with all applicable regulations.

8.3   Procedure for Sample Digestate Storage

      Sample digestates must be stored until 365 days after delivery of a
      complete, reconciled data package  to USEPA.

8.4   Contract Required Holding Time

      The maximum holding time for metals is 180 days from Validated Time of
      Sample Receipt  (VTSR).
                              Appendix B-ll                              ILM05.2

-------
Appendix B — Section 9
Calibration and Standardization

9.0   CALIBRATION AND STANDARDIZATION

9.1   Instrument Operating Conditions

      Because of the differences between various makes and models of
      satisfactory instruments, no detailed operating instructions can be
      provided.  Instead, the analyst should follow the instructions provided
      by the manufacturer of the particular instrument.  The Method Detection
      Limit  (MDL), precision, and interference effects must be investigated
      and established for each individual analyte line on that particular
      instrument.  All measurements must be within the instrument calibrated
      range.  It is- the responsibility of the analyst to verify that the
      instrument configuration and operating conditions used satisfy the
      analytical requirements and to maintain Quality Control  (QC) data
      confirming instrument performance and analytical results.

9.2   Graphite Furnace Atomic Absorption  (GFAA) Instrument Calibration
      Procedure

9.2.1    Instruments shall be calibrated daily or once every 24 hours and each
         time the instrument is set up.  The instrument standardization date
         and time shall be included in the raw data.

9.2.2    Calibration standards shall be prepared fresh daily or each time an
         analysis is to be made and discarded after use.  Prepare a blank and
         at  least three calibration standards in graduated amounts in the
         appropriate range.  One atomic absorption calibration standard shall
         be  at the CRQL.  The calibration standards shall be prepared using
         the same type of acid or combination of acids and at  the same
         concentration as will result in  the samples following sample
         preparation.

9.2.3    Calibration standards are prepared by diluting the stock metal
         solutions at the time of analysis.  Date and time of  preparation and
         analysis shall be given in the raw data.
 ILM05.2                       Appendix B-12

-------
                                                       Appendix B — Section 10
                                                                       Procedure

10.0  PROCEDURE

10.1  Sample Preparation

10.1.1   If insufficient sample amount  (less than 90% of the required amount)
         is received to perform the analyses, the Contractor shall contact
         Sample Management Office  (SMO) to inform them of the problem.  SMO
         will contact the Region for instructions.  The Region will either
         require that no sample analyses be performed or will require that a
         reduced volume be used for the sample analysis.  No other changes in
         the analyses will be permitted.  The Contractor shall document the
         Region's decision in the  Sample Delivery Group  (SDG) Narrative.

10.1.2   If multiphase samples (e.g., two-phase liquid sample, oily
         sludge/sandy soil sample) are received by the Contractor, the
         Contractor shall contact  SMO to apprise them of the type of sample
         received.  SMO will contact the Region.  If all phases of the sample
         are amenable to analysis, the Region may require the Contractor to do
         any of the following:

         00    Mix the sample and  analyze an aliquot from the homogenized
               sample.

         '-*1    Separate the phases of  the sample and analyze one or more of
               the  phases, separately.  SMO will provide EPA sample numbers
               for the additional  phases, if required.

         =*=    Do not analyze the  sample.

10.'1.2.1    If all of the phases are not amenable to analysis (i.e., outside
            scope), the Region may require the Contractor to do any of the
            following:

            '•*'•    Separate  the phases  and analyze  the  phase (s)  that is (are)
                  amenable  to analysis.  SMO will  provide EPA sample  numbers
                  for the additional phases, if  required.

            '•**    Do not analyze  the sample.

10.1.2.2    No other changes in the analyses will be permitted.  The
            Contractor shall document  the Region's decision in  the SDG
            Narrative.

10.1.3   Water/Aqueous Sample Preparation

10.1.3.1    Shake sample and transfer  50-100 mL of well-mixed sample to a 250
            mL heating vessel, add 1 milliliter  (mL) of  (1+1) HNO  3 and 2 mL of
            30% H202  to  the  sample.  Cover with watch glass or similar cover
            and heat on a hot plate, block digester, or  equivalent heating
            source which is adjustable and capable of  maintaining a
            temperature of  92-95°C for 2 hours or  until  sample  volume  is
            reduced to between 25  and  50 mL, making certain sample does not
            boil.  Cool sample and filter to remove insoluble material.

            NOTE: In place  of filtering, the sample, after  dilution and
            mixing, may be  centrifuged or allowed  to settle by  gravity
            overnight to remove insoluble material.

            Adjust sample volume  to  50-100 mL  with reagent  water.  The  sample
            is now ready for analysis.  Concentrations so determined  shall be
            reported as "total".   If volumes less  than 100  mL are used, all
            other reagents  shall  be  reduced appropriately  (e.g.,  if 50 mL is


                              Appendix B-13                             ILM05.2

-------
Appendix B — Section 10
Procedure (Con't)

            used, reduce reagent volumes by one-half).  The final volume of
            the  digestate must equal the initial volume of the sample aliquot.

10.1.4   Soil/Sediment Sample Preparation

10.1.4.1    A representative 1.0 gram  (g)  (wet weight) sample is digested in
            nitric acid and hydrogen peroxide.  The digestate is then refluxed
            with either nitric acid or hydrochloric acid.  Nitric acid  is
            employed as the final reflux acid for the Graphite Furnace  Atomic
            Absorption  (GFAA) analysis of  As, Pb, Se, and Tl.  A separate
            sample shall be dried for a percent solids determination.

10.1.4.2    Mix  the sample thoroughly to achieve homogeneity.  For each
            digestion procedure, weigh  (to the nearest 0.01 g) a 1.0 to 1.5 g
            portion of sample and transfer to a beaker.

10.1.4.3    Add  10 mL of 1:1 nitric acid  (HN03 ),  mix the slurry,  and cover
            with a watch glass.  Heat the  sample  to  92-95°C on hot plate or
            block digester, and reflux for 10 minutes without boiling.   Allow
            the  sample to cool, add 5 mL of concentrated  HNO 3, replace the
            watch glass, as appropriate, and reflux  for  30 minutes.  Do not
            allow the volume to be reduced to less than  5 mL while maintaining
            a  covering of solution over the bottom of the heating vessel.

10.1.4.4    After the second reflux step has been completed and the sample has
            cooled, add 2 mL of reagent water and 3 mL of 30% hydrogen
            peroxide  (H2O2) .  Return the heating vessel to the heat source for
            warming to start the  peroxide reaction.  Care must be taken to
            ensure that losses do not occur due to excessively vigorous
            effervescence.  Heat until effervescence  subsides, and cool the
            heat vessel.

            Continue to add 30% H 202 in 1  mL aliquots with warming until the
            effervescence is minimal or until the general sample appearance is
            unchanged.

            NOTE: Do not add more than a  total of 10  mL  30% H  202.

10.1.4.5    If the sample is being prepared for the GFAA  analysis of As, Pb,
            Se,  and Tl, continue heating  the acid-peroxide digestate until the
            volume has been reduced to approximately 2 mL, add 10 mL of
            reagent water,  and warm the mixture.  After  cooling,  filter
            through Whatman No. 42 filter  paper  (or  equivalent)  and dilute to
            100  mL with reagent water.

            NOTE: In place  of filtering,  the  sample  (after dilution and
            mixing) may be  centrifuged or  allowed to settle by gravity
            overnight  to remove insoluble  material.

            Dilute the  digestate  1:1  (200  mL  final volume) with  acidified
            water to maintain constant acid  strength.   For analysis, withdraw
            aliquots of appropriate volume,  and  add  any required reagent or
            matrix modifier.  The sample  is  now  ready for analysis.

10.2   Sample Analysis

10.2.1    Set up  instrument  with proper operating  parameters established in
          Section 9.1.

10.2.2    Profile and calibrate instrument  according  to  instrument
          manufacturer's recommended procedures,  using calibration  standard
          solutions.


ILM05.2                       Appendix B-14

-------
                                                       Appendix B — Section 10
                                                              Procedure  (Con't)


10.2.3   Instrument Parameters - Suggested Conditions

10.2.3.1    Arsenic

10.2.3.1.1     Wavelength: 193.7 run

10.2.3.1.2     Operating parameters should be set as specified by the
               particular instrument manufacturer.

10.2.3.1.3     The use of background correction is required. Background
               correction made by the deuterium arc method  does not adequately
               compensate for high levels of certain interferents  (i.e., Al,
               Fe).  If conditions occur where significant  interference  is
               suspected, the laboratory must switch to an  alternate
               wavelength or take other appropriate actions to compensate for
               the interference effects.

10.2.3.1.4     The use of the Electrodeless Discharge Lamps (EDLs) for the
               light source is recommended.

10.2.3.2    Lead

10.2.3.2.1     Wavelength: 283.3 nm

10.2.3.2.2     Operating parameters should be set as specified by the
               particular instrument manufacturer.

10.2.3.2.3     The use of background correction is required.

10.2.3.2.4     Greater sensitivity can be achieved using the 217.0 nm line,
               but the optimum concentration range is reduced.  The use  of an
               EDL at this lower wavelength has been found  to be  advantageous.
               Also a lower atomization temperature  (2400°C) may  be preferred.

10.2.3.2.5     To suppress sulfate interference  (up to 1500 ppm), lanthanum is
               added as the nitrate to both samples and calibration standards.

10.2.3.2.6     Since glassware contamination is a severe problem in lead
               analysis,  all glassware should be cleaned immediately prior to
               use, and once cleaned, should not be open to the atmosphere
               except when necessary.

10.2.3.3    Selenium

10.2.3.3.1     Wavelength: 196.0 nm

10.2.3.3.2     Operating parameters should be set as specified by the
               particular instrument manufacturer.

10.2.3.3.3     Selenium analysis suffers interference from  chlorides (>800
               mg/L) and  sulfate  (>200 mg/L).  For the analysis of industrial
               effluents  and samples with concentrations of sulfate from 200
               to  2000 mg/L, both samples and standards should be prepared to
               contain 1% nickel.

10.2.3.3.4     The use of the EDL for the light source is recommended.

10.2.3.4    Thallium

10.2.3.4.1     Wavelength: 276.8 nm
                              Appendix B-15                              ILM05.2

-------
Appendix B — Sections 10 & 11
Data Analysis and Calculations

10.2.3.4.2     Operating parameters should be set as specified by the
               particular instrument manufacturer.

10.2.3.4.3     The use of background correction is required.

10.2.3.4.4     Nitrogen may also be used as the purge gas.

11.0  DATA ANALYSIS AND CALCULATIONS

11.1  Water/Aqueous Sample Calculation

      The concentrations determined in the digestate are to be reported in
      units of microgram per Liter  ( :
      EQ. 1  Aqueous Sample Concentration

                                                 Vf
                             Concentration = C x — -
      WHERE,           C   =  Instrument value in ug/L

                       VŁ  =  Final digestion volume

                       Vt  =  Initial digestion volume

11.2  Soil Sample Calculation

      The concentrations determined in the digestate are to be reported on the
      basis of the dry weight of the sample, in units of milligram per
      kilogram  (mg/kg):
      EQ. 2  Soil Sample Concentration
                     Concentration (drywt.)  (mg/kg)
                               C x V
                               W x S
       WHERE,
C  =  Concentration (mg/L)

V  =  Final volume in liters after sample preparation

W  =  Weight in kg of wet sample

S  =  % Solids/100 (see Exhibit D - Introduction to
      Analytical Methods, Section 1.6).
 ILM05.2
       Appendix B-16

-------
                                                   Appendix B — Sections 11-17
                                                                Quality Control
11.3  Corrections For Sample Dilutions
      If dilutions were performed, the appropriate factor shall be applied to
      the sample values as follows:

      EQ. 3  Correction for Dilution

                               C  (ug/L) = Cl x DF
      WHERE,           C    = Concentration of analyte in sample

                       Cą   = Instrument value concentration

                       DF   = Dilution Factor

12.0  QUALITY CONTROL

      For specific Quality Control  (QC) requirements, the Contractor shall
      follow the instructions provided by the USEPA Region requesting the
      analysis.

13.0  METHOD PERFORMANCE

      Not applicable.

14.0  POLLUTION PREVENTION

      See Section 1.15 in Exhibit D - Introduction to Analytical Methods.

15.0  WASTE MANAGEMENT

      See Section 1.16 in Exhibit D - Introduction to Analytical Methods.

16.0  REFERENCES

16.1  US Environmental Protection Agency. Methods for Chemical Analysis of
      Water and Wastes. Method 206.2. March 1983.

16.2  US Environmental Protection Agency. Methods for Chemical Analysis of
      Water and Wastes. Method 239.2. March 1983.

16.3  US Environmental Protection Agency. Methods for Chemical Analysis of
      Water and Wastes. Method 270.2. March 1983.

16.4  US Environmental Protection Agency. Methods for Chemical Analysis of
      Water and Wastes. Method 279.2. March 1983.

17.0  TABLES/DIAGRAMS/FLOWCHARTS

      Not applicable.
                              Appendix B-17                             ILM05.2

-------
  EXHIBIT  B




INORGANIC FORMS
                                         ILM05.2

-------
                                 USEPA  - CLP

                                  COVER PAGE

Lab Name:  	   Contract:
Lab Code:  	 Case No. :  	 NRAS No. :  	 SDG No. :

SOW No.:  	

                EPA Sample No.                         Lab Sample ID
                                                             ICP-AES  ICP-MS
 Were ICP-AES and ICP-MS interelement              (Yes/No)
 corrections applied?

 Were ICP-AES and ICP-MS background corrections    (Yes/No)
 applied?

    If yes, were raw data generated before         (Yes/No)
    application of background corrections?

Comments:
I certify that this data package is in compliance with the terms and
conditions of the contract, both technically and for completeness,  for other
than the conditions detailed above.  Release of the data contained  in this
hardcopy data package and in the computer-readable data submitted on diskette
 (or via an alternate means of electronic transmission, if approved  in advance
by USEPA) has been authorized by the Laboratory Manager or the Manager's
designee, as verified by the following signature.

Signature:	  Name:	
Date:	              Title:	
                                  COVER  PAGE                            ILM05.2

-------
                                  USEPA - CLP
                                    1A-IN
                         INORGANIC ANALYSIS DATA SHEET
                                             EPA SAMPLE NO.
Lab Name:

Lab Code:
                 Contract:
Case No.:
NRAS No.:
                                        SDG NO.:
Matrix: (soil/water)

Level:  (low/med) 	

% Solids:  	
                       Lab Sample ID:

                       Date Received:
Concentration Units  (ug/L or mg/kg dry weight):
CAS No.
7429-90-5
7440-36-0
7440-38-2
7440-39-3
7440-41-7
7440-43-9
7440-70-2
7440-47-3
7440-48-4
7440-50-8
7439-89-6
7439-92-1
7439-95-4
7439-96-5
7439-97-6
7440-02-0
7440-09-7
7782-49-2
7440-22-4
7440-23-5
7440-28-0
7440-62-2
7440-66-6
57-12-5


Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide


Concentration


























C


























Q


























M


























Color Before:

Color After:

Comment s:
        Clarity Before:

        Clarity After:
                  Texture:
                  Artifacts:
                                   FORM IA-IN
                                                     ILM05.2

-------
Lab Name:

Lab Code:
                                  USEPA -  CLP

                                    IB-IN
                         INORGANIC ANALYSIS DATA SHEET
                 Contract:
Case No. -.
Matrix: (soil/water)

Level: (low/med) 	

% Solids:	
NRAS No.:
                       Lab Sample ID:

                       Date Received:
                                             EPA SAMPLE NO.
SDG No.:
Concentr
Color Be
Color Af
Comments
ation Units (ug/L or mg/kg dry weight]
CAS No.


























Analyte


























Cone ent rat i on


























.

C


























Q



























M


























fore : Claritv Before : Texture :

ter: Clarity After: Artifacts:




                                   FORM IB-IN
                                                     ILM05.2

-------
                                  USEPA -  CLP

                                     2 A-IN
                INITIAL AND CONTINUING  CALIBRATION VERIFICATION
Lab Name:

Lab Code:
                           Contract:
Case No.:
NRAS No.:
                                                           SDG No.:
Initial Calibration Verification Source:
Continuing Calibration Verification Source:

Concentration Units: ug/L

Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium -
Manganese -
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Initial Calibration
Verification
True




























Found




























%R(D




























Continuing Calibration Verification
True




























Found




























%R(D




























Found




























%R(1)




























M




























 (1)  Control Limits: Mercury 80-120; Other Metals  90-110; Cyanide  85-115
                                  FORM IIA-IN
                                                     ILM05.2

-------
                                 USEPA -  CLP

                                     2B-IN
                              CRQL  CHECK STANDARD
Lab Name:

Lab Code:
                           Contract:
Case No.:
NRAS No.:
SDG No.
CRQL Check Standard Source:

Concentration Units: ug/L
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide



CRQL Check Standard
Initial Final
True



























Found*



























%R(1)



























Found*



























%R(1)



























    (1) Control Limits: 70-130 with  the  following  exceptions:
    ICP-AES - Antimony, Lead, and Thallium:  50-150.
    ICP-MS - Cobalt, Manganese, and  Zinc:  50-150.

    *  If applicable, enter the concentration qualifier  "J"  or  "U"  after the
concentration in these columns  (e.g., 0.20U for Mercury).
                                  FORM IIB-IN
                                                     ILM05.2

-------
                                  USEPA -  CLP

                                     3-IN
                                    BLANKS
Lab Name:

Lab Code:
                           Contract:
Case No,:
NRAS No.:
                                                           SDG No.
Preparation Blank Matrix  (soil/water):
Preparation Blank Concentration Units  (ug/L or mg/kg):
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Initial
Calibration
Blank (ug/L)
















•












C




























Continuing Calibration
Blank (ug/L)
1




























C




























2




























C




























3




























C




























Preparation
Blank





























C





























M




























                                  FORM III-IN
                                                     ILM05.2

-------
                                 USEPA -  CLP
                       ICP-AES
                  4 A-IN
            INTERFERENCE  CHECK SAMPLE
Lab Name:

Lab Code:
                           Contract:
Case No.:
NRAS No.:
SDG No.:
ICP-AES Instrument ID:
                                   ICS Source:
Concentration Units:  ug/L
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc




True
Sol. Sol.
A AB




















































Initial Found
Sol. Sol.
A %R AB %R








































































































Final Found
Sol. Sol.
A %R AB %R








































































































                                  FORM IVA-IN
                                                     ILM05.2

-------
                                  USEPA -  CLP

                                     4B-IN
                       ICP-MS INTERFERENCE CHECK SAMPLE
Lab Name:

Lab Code:
                           Contract:
Case No.:
NRAS NO.:
                                                           SDG No.:
ICP-MS Instrument ID:
                     ICS Source:
Concentration Units: ug/L
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Carbon
Chloride
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Molybdenum
Nickel
Phosphorus
Potassium
Selenium
Silver
Sodium
Sulfur
Thallium
Titanium
Vanadium
Zinc




True
Sol. Sol.
A AB
































































Found
Sol . Sol .
A %R AB %R
































































































































                                  FORM IVB-IN
                                                                        ILM05.2

-------
Lab Name:

Lab Code:
                                  USEPA - CLP

                                     5A-IN
                         MATRIX SPIKE  SAMPLE RECOVERY
                 Contract:
                                             EPA SAMPLE NO.
Case No.:
NRAS No.:
SDG NO.:
Matrix: (soil/water)

% Solids for Sample:
                                    Level: (low/med)
Concentration Units  (ug/L or mg/kg dry weight):
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Control
Limit
%R




























Spiked Sample
Result (SSR)
C
























































Sample
Result (SR)
C
























































Spike
Added (SA)




























%R




























Q




























M




























Comments:
                                   FORM VA-IN
                                                     ILM05.2

-------
                                  USEPA - CLP
                                     5B-IN
                     POST-DIGESTION SPIKE SAMPLE RECOVERY
                                             EPA SAMPLE NO.
Lab Name:

Lab Code:
                 Contract:
Case No.:
NRAS No.:
SDG No.:
Matrix:  (soil/water)
                                    Level:  (low/med)
Concentration Units: ug/L
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Control
Limit
l_ %R
















„










Spiked Sample
Result (SSR)
C






















































Sample
Result (SR)
C






















































Spike
Added (SA)



























%R



























Q



























M



























Comment s:
                                   FORM VB-IN
                                                                        ILM05.2

-------
                                  USEPA -  CLP
Lab Name:

Lab Code:
Case No.:
                                     6-IN
                                  DUPLICATES
Contract:  	

  NRAS No.:
                                             EPA SAMPLE NO.
SDG NO.:
Matrix: (soil/water)

% Solids for Sample:
                                    Leve1:  (1ow/med)

                             %  Solids  for Duplicate:
Concentration Units  (ug/L or mg/kg dry weight):
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Control
Limit




























Sample (S)
C









































-














Duplicate (D)
C
























































RPD




























Q




























M




























                                   FORM VI-IN
                                                                        ILM05.2

-------
                                  USEPA - CLP

                                     7-IN
                           LABORATORY CONTROL SAMPLE
Lab Name:

Lab Code:
                           Contract:
Case No.:
NRAS No.:
SDG No.
Solid LCS Source:
Aqueous LCS Source:
Analyte
Aluminum
Ant imony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Aqueous (ug/L)
True Found %R




















































































Solid (mg/kg)
True Found C Limits %R








































































































































































                                  FORM VII-IN
                                                     ILM05.2

-------
                                  USEPA - CLP
Lab Name:

Lab Code:
                                     8-IN
                      ICP-AES and ICP-MS SERIAL DILUTIONS
                 Contract:
                                             EPA SAMPLE NO.
Case No.
Matrix: (soil/water)
NRAS No.:
                                                            SDG No.:
                                   Level:  (low/med)
Concentration Units: ug/L
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc




Initial Sample
Result (I)
C




















































Serial
Dilution
Result (S)
C




















































%
Difference


























Q


























M


























                                  FORM VIII-IN
                                                     ILM05.2

-------
                                  USEPA -  CLP

                                     9-IN
                      METHOD DETECTION LIMITS  (ANNUALLY)
Lab Name:

Lab Code:
                           Contract:
Case No.:
NRAS No.:
Instrument Type:
            Instrument ID:
SDG No.

 Date:
Preparation Method:
Concentration Units  (ug/L or mg/kg):
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide




Wavelength
/Mass




























CRQL




























MDL




























Comments:
                                   FORM IX-IN
                                                                        ILM05.2

-------
                                  USEPA - CLP

                                    10A-IN
              ICP-AES INTERELEMENT CORRECTION FACTORS  (QUARTERLY)
Lab Name:

Lab Code:
                           Contract:
Case No.:
ICP-AES Instrument ID:
NRAS No.:

  Date:
SDG NO.
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc




Wave-
length
(nm)


























Interelement Correction Factors for:
Al Ca Fe Mg


































































































































Comments:
                                   FORM XA-IN
                                                     ILM05.2

-------
                                  USEPA -  CLP

                                    10B-IN
              ICP-AES INTERELEMENT CORRECTION FACTORS (QUARTERLY)
Lab Name:

Lab Code:
                           Contract:
Case No.:
ICP-AES Instrument ID:
                   NRAS No.:

                     Date :
                                                           SDG No.
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc




Wave-
length
(nm)


























Interelement Correction Factors for:


































































































































Comments:
                                   FORM XB-IN
                                                                        ILM05.2

-------
                                  USEPA - CLP

                                     11-IN
                 ICP-AES and ICP-MS LINEAR RANGES  (QUARTERLY)
Lab Name:

Lab Code:
Case No.:
ICP Instrument ID:
                           Contract:
NRAS No.:

  Date .-
SDG No.
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc




Integ.
Time
(Sec.)


























Concentration
(ug/L)


























M


























Comments:
                                   FORM XI-IN
                                                     ILM05.2

-------
 USEPA  -  CLP

     12-IN
PREPARATION LOG
Lab Name :
Lab Code : C
Preparation Method:

EPA
Sample
No.


































'ase No. : N


Preparation
Date

































Contract :
RAS No . :


Weight
(gram)


































SDG No . :


Volume
(mL)

































  FORM XII-IN
                                        ILM05.2

-------
                                  USEPA - CLP

                                     13-IN
                               ANALYSIS RUN LOG
Lab Name:

Lab Code:
                           Contract:
Case No.:
NRAS No.:
SDG NO.
Instrument ID:

Start Date:
                         Analysis Method:

                         End Date: 	
EPA
Sample
No.




































D/F





































Time




































Analytes
A
L




































S
B




































A
S




































B
A




































B
E




































C
D




































C
A




































C
R




































C
O




































C
U




































F
E




































P
B




































M
G




































M
N




































H
G




































N
I




































K




































S
E





















^














A
G




































N
A




































T
L




































V




































Z
N




































C
N














































































































                                  FORM XIII-IN
                                                     ILM05.2

-------
                                  USEPA - CLP

                                     14-IN
                                  TCP-MS Tune
Lab Name:

Lab Code:
Case No.:
ICP-MS Instrument ID:
                           Contract:
NRAS No.:
                                        SDG NO.

                                         Date:
Element - Mass
Be - 9
Mg - 24
Mg - 25
Mg - 26
Co - 59
In - 113
In - 115
Pb - 206
Pb - 207
Pb - 208




Avg. Measured Mass (amu)














Avg. Peak Width at
5% Peak Height (amu)














%RSD














Comments:
                                  FORM XIV-IN
                                                     ILM05.2

-------
                                 USEPA -  CLP

                                     15-IN
             ICP-MS Internal Standards Relative Intensity Summary
Lab Name:

Lab Code:
                           Contract:
Case No.:
NRAS No.:
ICP-MS Instrument ID:
                     Start Date:
SDG No.:  _

  End Date:
EPA Sample
No.
























Time























Internal Standards %RI For:
Element























Q























Element























Q























Element























Q























Element























Q























Element























Q























                                   FORM XV-IN
                                                     ILM05.2

-------
                                          SAMPLE LOG-IN SHEET
ab Name
eceived By (Print Name)
Page 	 of 	
Log- in Date
eceived By (Signature)
ase Number
emarks :
. Custody Seal(s)
. Custody Seal Nos .
Present /Absent*
Intact /Broken

. Traffic Present/Absent*
Reports /Chain of
Custody Records or
Packing Lists
. Airbill
. Airbill No.
Sample Tags
Airbill/Sticker
Present /Absent*

Present /Absent*
Sample Tag Numbers Listed/Not
Listed on
Traffic
Report /Chain of
Custody Record
Sample Condition
Intact /Broken*/
Leaking
. Cooler Temperature Present/Absent*
Indicator Bottle
Cooler Temperature
0. Does information
on Traffic
Reports/Chain of
Custody Records
and sample tags
agree?
1 . Date Received at
Lab
2 . Time Received
Yes/No*

Sample Transfer
raction
rea #
y
n
Fraction
Area #
By
On
Sample Delivery Group No.
EPA
Sample #



















Aqueous
Sample pH


















Cor re spending
Sample Tag #


















Assigned
Lab S


















NRAS Number

Remarks :
Condition of
Sample
Shipment, etc.


















 Contact SMO and attach record of resolution
eviewed By
ate


Logbook No.
Logbook Page No.
                                               FORM  DC-I
ILM05.2

-------
                    FULL INORGANICS COMPLETE SDG FILE (CSF)  INVENTORY SHEET
     LABORATORY NAME

     CITY/STATE 	
     CASE  NO.  	 SDG NO.

     SDG NOs.  TO FOLLOW 	

     NRAS  NO.  	


     CONTRACT  NO.  	

     SOW NO. 	
     All  documents delivered in the Complete SDG File must be original documents
     where  possible.   (Reference - Exhibit B Section 2.6)

                                                 PAGE NOs.               CHECK
                                              FROM         TO         LAB      REGION
1.  Inventory Sheet (DC-2)   (Do not number)

2.  Sample  Log-In Sheet (DC-1)                	        	        	      	

3.  Traffic Report/Chain of Custody Record    	        	        	      	

4.  Cover Page                                	        	        	      	

5.  SDG Narrative                             	        	        	      	
    Inorganic Analysis
6.  Data Sheet (Form I-IN)

7.  Initial & Continuing Calibration
    Verification (Form IIA-IN)

8.  CRQL Standard
    (Form IIB-IN)

9.  Blanks (Form III-IN)

10.  ICP-AES Interference Check Sample
    (Form IVA-IN)

11.  ICP-MS Interference Check Sample
    (Form IVB-IN)

12.  Matrix Spike Sample Recovery
    (Form VA-IN)

13.  Post-Digestion Spike Sample Recovery
    (Form VB-IN)

14.  Duplicates (Form VI-IN)

15.  Laboratory Control Sample
    (Form VII-IN)

16.  ICP-AES and ICP-MS Serial Dilutions
    (Form VIII-IN)

17.  Method Detection Limits  (Annually)
    (Form IX-IN)


18.  ICP-AES Interelement Correction
    Factors  (Quarterly)  (Form XA-IN)


19.  ICP-AES Interelement Correction
    Factors  (Quarterly)  (Form XB-IN)

20.  ICP-AES and ICP-MS Linear Ranges
    (Quarterly)  (Form XI-IN)

21.  Preparation Log  (Form XII-IN)

22.  Analysis Run Log  (Form XIII-IN)
                                          FORM DC-2-1                            ILM05.2

-------
                                                 PAGE NOs.               CHECK
                                              FROM         TO         LAB     REGION
23. ICP-MS Tune (Form XIV-IN)

24. ICP-MS Internal Standards Relative
    Intensity Summary (Form XV-IN)
25. ICP-AES Raw Data

26. GFAA Raw Data (If Applicable)

27. ICP-MS Raw Data

28. Mercury Raw Data

29. Cyanide Raw Data

30. Preparation Logs Raw Data

31. Percent Solids Determination Log

32. USEPA Shipping/Receiving Documents
        Airbill  (No.  of  Shipments 	)

        Sample Tags

        Sample Log-In Sheet  (Lab)

33. Misc.  Shipping/Receiving Records
    (list all individual records)
        Telephone Logs
34. Internal Lab Sample Transfer Records &
    Tracking Sheets  (describe or list)
35. Internal Original Sample Prep &
    Analysis Records  (describe or list)
         Prep  Records  	
         Analysis Records

         Description 	
36. Other Records  (describe or list)
         Telephone Communications Log
37. Comments:
Completed by:
 (CLP Lab)
           (Signature)                         (Print  Name  &  Title)              (Date)
Audited by:
 (USEPA)    	   	   	
           (Signature)                         (Print  Name  &  Title)              (Date)

                                          FORM DC-2-2                           ILM05.2

-------