USEPA Contract Laboratory Program
Statement of Work for Organics Analysis
   Multi-media, Multi-concentration

-------
                            54OR995O7
USEPA CONTRACT LABORATORY PROGRAM



       STATEMENT OF WORK

             FOR

       ORGANICS ANALYSIS


Multi-Media,  Multi-Concentration

           OLM04.2
           May 1999

-------
                              STATEMENT  OF WORK

                              TABLE  OF CONTENTS



EXHIBIT A:  SUMMARY OF REQUIREMENTS

EXHIBIT B:  REPORTING AND DELIVERABLES REQUIREMENTS

EXHIBIT C:  TARGET COMPOUND LIST  (TCL) AND CONTRACT REQUIRED QUANTISATION
            LIMITS (CRQLs)

EXHIBIT D:  ANALYTICAL METHODS

EXHIBIT E:  QUALITY ASSURANCE/QUALITY CONTROL PROCEDURES AND REQUIREMENTS

EXHIBIT F:  CHAIN-OF-CUSTODY, DOCUMENT CONTROL, AND WRITTEN STANDARD OPERATING
            PROCEDURES

EXHIBIT G:  GLOSSARY OF TERMS

EXHIBIT H:  AGENCY STANDARD IMPLEMENTATION
                                                                        OLM04.2

-------
       EXHIBIT A
SUMMARY OF REQUIREMENTS
          A-l                               OLM04.2

-------
                     Exhibit A - Summary of Requirements

                              Table of Contents


Section                                                                   Page

1.0   PURPOSE	3

2.0   DESCRIPTION OF SERVICE	3

3.0   DATA USES	   3

4.0   SUMMARY OF REQUIREMENTS	3
      4 .1   Introduction to the SOW	3
      4.2   Overview of Major Task Areas	4
      4.3   Technical and Management Capability 	   9
                                      A-2                               OLM04.2

-------
                                                     Exhibit A -- Sections 1-4
                                                       Summary of Requirements
1.0   PURPOSE
      The purpose of the multi-media, multi-concentration organic analytical
      service is to provide analytical data for use by the U.S. Environmental
      Protection Agency (EPA) in support of its investigation and clean-up
      activities under the Comprehensive Environmental Response, Compensation,
      and Liability Act of 1980 (CERCLA) and the Superfund Amendments and
      Reauthorization Act of 1986  (SARA).   Other EPA Program Offices that have
      similar analytical data needs also use this service.

2.0   DESCRIPTION OF SERVICE

      The organic analytical service provides a contractual framework for
      laboratories to apply EPA Contract Laboratory Program (CLP) analytical
      methods for the isolation, detection, and quantitative measurement of 48
      volatile, 65 semivolatile, and 28 pesticide/Aroclor target compounds in
      water and soil/sediment environmental samples.  The analytical service
      provides the methods to be used, and the specific contractual
      requirements by which EPA will evaluate the data.  This service uses gas
      chromatography/mass spectrometry  (GC/MS) and gas chromatography/electron
      capture  (GC/EC) methods to analyze the target compounds.

3.0   DATA USES

      This analytical service provides data which EPA uses for a variety of
      purposes, such as determining the nature and extent of contamination at
      a hazardous waste site, assessing priorities for response based on risks
      to human health and the environment, determining appropriate cleanup
      actions, and determining when remedial actions are complete.  The data
      may be used in all stages in the investigation of a hazardous waste
      site, including site inspections, Hazard Ranking System scoring,
      remedial investigations/feasibility studies, remedial design,
      treatability studies, and removal actions.  In addition, this service
      provides data that are available for use in Superfund
      enforcement/litigation activities.

4.0   SUMMARY OF REQUIREMENTS

4.1   Introduction to the SOW.  This SOW is designed as part of the
      documentation for a contract between EPA and a commercial laboratory
      performing analyses in support of EPA Superfund programs.  The SOW is
      comprised of eight exhibits.  Exhibit A provides an overview of the SOW
      and its general requirements..  Exhibit B contains a description of the
      reporting and deliverables requirements, in addition to the data
      reporting forms and the forms instructions.  Exhibit C specifies the
      target compound list for this SOW with the contract-required
      quantitation limits for sample matrices.  Exhibit D details.the specific
      analytical procedures to be used with this SOW and resulting contracts.
      Exhibit E provides descriptions of required quality assurance/quality
      control  (QA/QC), standard operating procedures, and procedures used for
      evaluating analytical methodologies, QA/QC performance, and the
      reporting of data.  Exhibit F contains chain-of-custody -and sample
      documentation requirements which the Contractor shall follow.  To ensure
      proper understanding of the terms utilized in this SOW, a glossary can

                                      A-3                               OLMO'4.2

-------
Exhibit A -- Section 4
Summary of Requirements

      be found in Exhibit G (when a term is used in the text without
      explanation,  the glossary meaning shall be applicable).   Specifications
      for reporting data in computer-readable form appear in Exhibit H.

4.2   Overview of Major Task Areas.  For each sample,  the Contractor shall
      perform the tasks described in this section.  Specific requirements for
      each task are detailed in the exhibits as referenced.

4.2.1    Task I:  Chain-of-Custody

4.2.1.1     Chain-of-Custody.  The Contractor shall receive and maintain
            samples under proper chain-of-custody procedures.   All associated
            document control and inventory procedures shall be developed and
            followed.  Documentation, as described therein, shall be required
            to show that all procedures are being strictly followed.  This
            documentation shall be reported as the Complete Sample Delivery
            Group File  (CSF) (see Exhibit B).  The Contractor shall establish
            and use appropriate procedures to handle confidential information
            received from the Agency.  See Exhibit F for specific
            requirements.

4.2.1.2     Sample Scheduling/Shipments.  Sample shipments to the Contractor's
            facility will be scheduled and coordinated by the CLP Sample
            Management Office  (SMO).  The Contractor shall communicate with
            SMO personnel by telephone, as necessary throughout the .process of
            sample scheduling, shipment, analysis, and data reporting, to
            ensure that  samples are properly processed.  At the time of sample
            scheduling,  the Contractor will be notified if  the Modified SW-846
            Method 5035  is to be used in the preparation and analysis  of low
            level soil samples for volatiles.

4.2.1.2.1      Samples will be shipped routinely to the Contractor through an
               overnight delivery service.  However, as necessary, the
               Contractor shall be responsible for any handling or processing
               required  for the receipt of sample shipments, including pick-up
               of samples at the nearest servicing airport, bus station, or
               other carrier service within the Contractor's geographical
               area.  The Contractor shall be available to  receive sample
               shipments at any time the delivery service is operating,
               including Saturdays.

4.2.1.2.2      If there  are problems with the samples  (e.g., mixed media,
               containers broken or  leaking) or sample documentation/paperwork
                (e.g., Traffic  Reports not with shipment, sample and Traffic
               Report numbers  do not correspond), the Contractor shall
               immediately contact SMO  for  resolution.  The Contractor shall
               immediately notify  SMO regarding any problems and laboratory
               conditions that affect the timeliness of analyses and data
               reporting.   In  particular, the Contractor shall notify  SMO
               personnel in  advance  regarding sample data  that will be
               delivered late  and  shall specify the estimated delivery date.

4.2.1.2.3      To effectively  monitor the temperature of the sample shipping
               cooler,  each USEPA  Regional  office may include a sample

                                      A-4                              OLM04.2

-------
                                                        Exhibit A -- Section 4
                                                       Summary of Requirements

               shipping cooler temperature blank with each cooler shipped.
               The temperature blank will be clearly labeled:-  USEPA COOLER
               TEMPERATURE INDICATOR.

4.2.1.2.3.1       When the USEPA Regional office supplies a cooler temperature
                  indicator bottle in the sample shipping cooler, the
                  Contractor shall use the USEPA supplied cooler temperature
                  indicator bottle to determine the cooler temperature.   The
                  temperature of the cooler shall be measured at the time of
                  sample receipt by the Contractor.

4.2.1.2.3.2       The temperature of the sample shipping cooler-shall be
                  measured and recorded immediately upon opening the cooler,
                  and prior to unpacking the samples or removing the packing
                  material.

4.2.1.2.3.3       To determine the temperature of the cooler, the contractor
                  shall locate the cooler temperature indicator bottle in the
                  sample shipping cooler, remove the cap, and insert a
                  calibrated thermometer into the cooler temperature indicator
                  bottle.  Prior to recording the temperature, the Contractor
                  shall allow a minimum of 3 minutes, but not greater than 5
                  minutes, for the thermometer to equilibrate with the liquid
                  in the bottle.  At a minimum, the calibrated thermometer
                  (±1°C)  shall have a measurable range of 0 to 50 degrees
                  Celsius.  Other devices which can measure temperature may be
                  used if they can be calibrated to ±1°C and have a range of 0
                  to 20°C.  Jf a temperature indicator bottle is not present
                  in the cooler, an alternative means of determining cooler
                  temperature shall be used.  However, under no circumstances
                  shall a thermometer or any other device be inserted into a
                  sample bottle for the purpose of determining cooler
                  temperature.  The Contractor shall contact SMO and inform
                  them that a temperature indicator bottle was not present in
                  the cooler.  The Contractor shall document the alternative
                  technique used to determine cooler temperature in the SDG
                  Narrative.

4.2.1.2.3.4       If the temperature of the sample shipping cooler's
                  temperature indicator exceeds 10 degrees Celsius, the
                  Contractor shall contact SMO and inform them of the
                  temperature deviation.  The SMO will contact the Region from
                  which the samples were shipped for instructions on how to
                  proceed.  The Region will either require that no sample
                  analysis(es) be performed or that the Contractor proceed
                  with the analysis(es).  The SMO will in turn notify the
                  Contractor of the Region's decision.  The Contractor shall
                  document the Region's decision in the SDG Narrative.  Also
                  in the SDG Narrative, the Contractor shall list by fraction,
                  the USEPA sample-number of all samples which were shipped in
                  a cooler which exceeded 10 degrees Celsius.
                                      A-5                              OLM04.2

-------
Exhibit A -- Section 4
Summary of Requirements

4.2.1.2.3.5       The Contractor shall record the temperature of the cooler on
                  the DC-1 Form, under Remark #9 - Cooler Temperature, and in
                  the SDG Narrative.

4.2.1.2.4   .   The Contractor shall accept all samples scheduled by SMO,
               provided that the total number of samples received in any
               calendar month does not exceed the monthly limitation expressed
               in the contract.   Should the Contractor elect to accept
               additional samples, the Contractor shall remain bound by all
               contract requirements for analysis of those samples accepted.

4.2.1.2.5      The Contractor shall be required to routinely return sample
               shipping containers  (e.g., coolers) to the appropriate sampling
               office within 14 calendar days.following shipment receipt '(see
               Clause entitled Government Furnished Supplies and Materials).

4.2.2    Task II:  Analysis of Samples

4.2.2.1     Overview.  Sample.analyses will be scheduled by groups of samples,
            each defined as a Case and identified by a unique EPA Case number
            assigned by SMO.  A Case signifies a group of samples collected at
            one site or geographical area over a finite time period, and will
            include one or more field samples with associated blanks.  Samples
            may be shipped to the Contractor  in a single shipment or multiple
            shipments over a period of time,  depending on the size of the
            Case.

4.2.2.1.1      A Case consists of one or more Sample Delivery Group(s).  A
               Sample Delivery Group  (SDG)  is defined by the following,
               whichever is most  frequent:

               •   Each Case of field samples received, OR

               •   Each 20 field  samples  (excluding PE samples) within  a Case,
                   OR

               •   Each  7 calendar  day period (excluding Sundays and
                   Government holidays)  during which field  samples  in a Case
                   are received (said period  beginning with the receipt of  the
                   first sample in  the SDG).

               In  addition, all samples  and sample fractions assigned to an
               SDG must have  been scheduled under the same  contractual
               turnaround time.

 4.2.2.1.2      Samples may be assigned to SDGs by matrix  (i.e., all  soils  in
               one SDG,  all waters  in another),  at the discretion of the
               laboratory.  However,  PE  samples  received within a Case  shall
               be  assigned  to an  SDG containing field samples  for that  Case.
               Such  assignment  shall be  made  at  the time the samples are
               received,  and  shall  not be made  retroactively.

 4.2.2.2     Preparation  Techniques.  The Contractor will prepare samples as
             described in Exhibit  D.   For semivolatile and pesticide/Aroclor

                                      A-6                               OLM04.2

-------
                                                        Exhibit A -- Section 4
                                                       Summary of Requirements

            samples, an aliquot is extracted with a solvent and concentrated.
            The concentrated extract is subjected to fraction-specific cleanup
            procedures and then analyzed by GC/MS for semivolatile or GC/EC
            for the pesticide/Aroclor target compounds listed in Exhibit C.
            For volatile samples,  an aliquot is purged with an inert gas,
            trapped on a solid sorbent, and then desorbed onto the GC/MS for
            analysis of the target compounds listed in Exhibit C.

4.2.2.3     Analytical Techniques.  The target compounds listed in Exhibit C
            shall be identified as described in the methodologies given in
            Exhibit D.  Automated computer programs may be used to facilitate
            the identification of compounds.

4.2.2.4     Qualitative Verification of Compounds.  The volatile and
            semivolatile compounds identified by GC/MS techniques shall be
            verified by an analyst competent in the interpretation of mass
            spectra by comparison of the suspect mass spectrum to the mass
            spectrum of a standard of the suspected compound.  This procedure
            requires the use of multiple internal standards.

4.2.2.4.1      If a compound initially identified by GC/MS techniques cannot
               be verified, but in the technical judgment of the mass spectral
               interpretation specialist the identification is correct, then
               the Contractor shall report that identification and proceed
               with quantitation.

4.2.2.4.2      The pesticide/Aroclor compounds identified by GC/EC techniques
               shall be verified by an analyst competent in the interpretation
               of gas chromatograms and by comparison of the retention times
               of the suspected unknowns with the retention times of
               respective standards of the suspected compounds.  Compounds
               shall also be confirmed by GC/MS techniques if the compounds
               are of sufficient concentration to be detected by the GC/MS.

4.2.2.5     Quantitation of Verified Compounds.  The Contractor shall
            quantitate components identified by GC/MS techniques by the
            internal standard method stipulated in Exhibit D.  Where multiple
            internal standards are required by EPA, the Contractor shall
            perform quantitation utilizing the internal standards specified in
            Exhibit D.  The Contractor shall quantitate components analyzed -by
            GC/EC techniques by the external standard method stipulated in
            Exhibit D.  The Contractor shall also perform an initial three-
            point calibration, verify its linearity, determine the breakdown
            of labile components,  and determine calibration factors for all
            standards analyzed by GC/EC techniques as described in Exhibit D.

4.2.2.6     Tentative Identification of Non-Target Sample Components.  For
            each analysis of a sample, the Contractor shall conduct mass
            spectral library searches to determine tentative compound
            identifications as follows:  For each volatile sample, the
            Contractor shall'conduct a search to determine the possible
            identity of up to 30 organic compounds of greatest concentration
            .which are not system monitoring compounds or internal standards
            and are not listed in Exhibit C under volatiles or semivolatiles.

                                      A-7                              OLM04.2

-------
Exhibit A -- Section 4
Summary of Requirements

            For each semivolatile sample,  the Contractor shall conduct a
            search to determine the possible identification of up to 30
            organic compounds of greatest concentration which are not
            surrogates or internal standards and are not listed in Exhibit C
            under volatiles or semivolatiles.  In performing searches, the
            NIST/EPA/NIH (May 1992 release or later)  and/or Wiley (1991
            release or later),  or equivalent, mass spectral library shall be
            used.

            NOTE:  Substances with responses less than 10% of the nearest
            internal standard are not required to be searched in this fashion.

4.2.2.7     Quality Assurance/Quality Control Procedures.  The Contractor
            shall strictly adhere to all specific QA/QC procedures prescribed
            in Exhibits D and E.  Records documenting the use of the protocol
            shall be maintained in accordance with the document control
            procedures prescribed in Exhibit F, and shall be reported in
            accordance with Exhibit B and Exhibit H.

4.2.2.7.1      The Contractor shall maintain a Quality Assurance Plan  (QAP)
               with the objective of providing sound analytical chemical
               measurements.  This program shall incorporate the quality
               control procedures, any necessary corrective action, and all
               documentation required during data collection as well as the
               quality assessment measures performed by management to ensure
               acceptable data production.

4.2.2.7.2      Additional quality control shall be conducted in the form of
               the analysis of laboratory evaluation samples submitted to  the
               laboratory by the Agency.  The results of all such quality
               control or laboratory evaluation samples may be used as the
               basis  for an equitable adjustment to reflect the reduced value
               of the data to the Agency or  rejection of data for:  sample(s),
               a fraction within an SDG, or  the entire SDG, and/or may be  used
               as the basis  for contract action.  "Compliant performance"  is
               defined as that which yields  correct compound identification
               and concentration values as determined by the Agency, as well
               as meeting the contract requirements for analysis  (Exhibit  D),
               quality assurance/quality control  (Exhibit E), data reporting
               and other deliverables  (Exhibits B and H) , and sample custody,
               sample documentation, and standard operating procedure
               documentation (Exhibit F).

4.2.2.8     The  Contractor may be requested  by EPA  to perform modified
             analyses.  These modifications may include, but  are not limited
             to,  additional compounds, sample matrices other  than  soil/sediment
             or water, and lower guantitation limits.  These  requests  will  be
             made by  the EPA  Administrative Project Officer and Contracting
             Officer  in writing, prior  to sample  scheduling.   If  the Contractor
             voluntarily elects  to perform these  modified analyses,  these
             analyses will be performed with  no increase  in per sample  price.
             In addition, all applicable  contract requirements specified in the
             Statement of Work/Specifications will remain in  effect.
                                      A-8                               OLM04.2

-------
                                                        Exhibit A -- Section 4
                                                       Summary of Requirements
4.2.3    Task III:  Reporting Requirements
4.2.3.1     EPA has provided the Contractor with formats for the reporting of
            data  (Exhibits B and H).   The Contractor shall be responsible for
            completing and submitting analysis data sheets and computer-
            readable data on diskette (or via an alternate means of electronic
            transmission approved in advance by the EPA) in the format
            specified in this SOW and within the time specified in the
            Contract Performance/Delivery•Schedule in Exhibit B.

4.2.3.2     Use of formats other than those designated by EPA will be deemed
            as noncompliant.   Such data are unacceptable.  Resubmission in the
            specified format at no additional cost to the Agency shall be
            required.

4.2.3.3     Computer-generated forms may be submitted in the hardcopy data
            package(s) provided that the forms are in exact EPA format.  This
            means that the order of data elements is the same as on each EPA-
            required form, including form numbers and titles, page numbers,
            and header information.

4.2.3.4     The data reported by the Contractor on the hardcopy data forms and
           • the associated computer-readable data submitted by the Contractor
            shall contain identical information.  If discrepancies are found
            during government inspection, the Contractor shall be required to
            resubmit either the hardcopy forms or the computer-readable data,
            or both sets of data, at no additional cost to the Agency.

4.3   Technical and Management Capability

4.3.1    Personnel.  The Contractor shall have adequate personnel at all times
         during the performance of the contract to ensure that EPA receives
         data that meet the terms and conditions of the contract.

4.3.2    Instrumentation.  The Contractor shall have sufficient gas
         chromatograph/electron capture/data system (GC/EC/DS), gas
         chromatograph/mass spectrometer/data system (GC/MS/DS), including
         magnetic tape storage devices,  and gel permeation chromatography
         system (GPC) capability to meet all the terms and conditions of the
         contract.

4.3.3    Facilities.   The Contractor shall maintain a facility suitable for
         the receipt, storage, analysis, and delivery of the product meeting
         the terms and conditions of the contract.
                                      A-9                               OLM04.2

-------
               EXHIBIT B
REPORTING AND DELIVERABLES REQUIREMENTS
                   B-l                               OLM04.2

-------
             Exhibit B  - Reporting  and  Deliverables Requirements

                              Table of  Contents
Section                                                                   Page

1.0   CONTRACT REPORTS/DELIVERABLES DISTRIBUTION  	  3
      1.1   Report Deliverable Schedule	3
      1.2   Distribution  	  6

2.0   REPORTING REQUIREMENTS AND ORDER OF DATA DELIVERABLES 	  7
      2.1   Introduction	7
      2.2   Resubmission of Data	7
      2.3   Quality Assurance Plan and Standard Operating Procedures  ...  8
      2.4   Sample Traffic Reports   	  8
      2 . 5   Sample Data Summary Package	9
      2.6   Sample Data Package	  9
      2.7   Complete SDG File	25
      2.8   Data in Computer-Readable Form	26
      2.9   Preliminary Results 	 27
      2.10  GC/MS and GC/EC Tapes	27
      2.11  Extracts	27

3.0   FORMS INSTRUCTIONS  	 28
      3.1   Introduction	28
      3.2   General Information 	 28
      3.3   Header Information	29
      3.4   Organic Analysis Data Sheet (Form I, All Fractions)	33
      3.5   Organic Analysis Data Sheet:  Tentatively Identified Compounds
            (Form I VOA-TIC and Form I SV-TIC)	 38
      3.6   System Monitoring Compound Recovery (Form II, VOA-1, VOA-2)  .   . 39
      3.7   Surrogate Recovery  (Form II, SV-1, SV-2 and Form II, PEST-1,
            PEST-2)	39
      3.8   Matrix Spike/Matrix Spike Duplicate Recovery  (Form III, All
            Fractions)   	41
      3.9   Method Blank Summary  (Form IV, All Fractions) 	 42
      3.10  GC/MS Instrument Performance Check and Mass Calibration
            (Form V VOA and Form V SV)  	'.	44
      3.11  GC/MS Initial Calibration Data  (Form VI, VOA-1, VOA-2 and
            Form VI, SV-1, SV-2)  	45
      3.12  GC/EC Initial Calibration Data  (Form VI, PEST-1, PEST-2)  ... 46
      3.13  GC/MS Continuing Calibration Data (Form VII, VOA-1, VOA-2
            and Form VII,  SV-1, SV-2)	48
      3.14  GC/EC Calibration Verification Summary  (Form VII, PEST-1,
            PEST-2)	49
      3.15  Internal Standard Area and RT Summary (Form VIII VOA and Form
            VIII, SV-1, SV-2)	50
      3.16  Pesticide Analytical Sequence (Form VIII PEST)  	 52
      3.17  Pesticide Cleanup Summary (Form IX,  PEST-1, PEST-2) 	 53
      3.18  Pesticide/Aroclor Identification  (Form X,  PEST-1, PEST-2) ... 54
      3.19  Sample Log-In Sheet (Form DC-1)  	 56
      3.20  Document Inventory Sheet  (Form DC-2)  	 57

4.0   DATA REPORTING FORMS	59
                                      B-2                              OLM04.2

-------
                                                          Exhibit B--Section 1
                                    Contract Reports/Deliverables Distribution

1.0   CONTRACT REPORTS/DELIVERABLES DISTRIBUTION

1.1   Report Deliverable Schedule.  The following table reiterates the
      contract reporting and deliverable requirements specified in the
      Contract Schedule  (Performance/Delivery Schedule) and specifies the
      distribution that is required for each deliverable.  The turnaround
      times for items B through E listed below are 7, 14, and 21 days.

      NOTE: Specific recipient names and addresses are subject to change
      during the term of the contract.  The Administrative Project Officer
      (APO) will notify the Contractor in writing of such changes when they
      occur.

                                    Table 1


                                                              Distribution



Item


No. of
Copies*



Delivery Schedule

o
• 2
W
O
•H
01
0)

 A.
   1. 3
  F.'
Sample Traffic
Reports
Sample Data
Summary
Package

Sample Data
Package0

Data in
Computer
Readable
Format

Complete SDG
File


Preliminary
Results  (VOA
Analyses)
            Preliminary
            Results (SV
            and Pest
            Analyses)
3 working days
after receipt of
last  sample in
Sample Delivery
Group (SDG).2
XXs days  after
receipt of last
sample in SDG.

XXs days  after
receipt of last
sample in SDG.

XXs days  after
receipt of last
sample in SDG.


XX3 days  after
receipt- of last
sample in SDG.

Within 48 hours
after receipt of
last sample  in SDG
at  laboratory, if
requested.

Within 72 hours
after receipt of
last sample  in SDG
at  laboratory, if
requested.
                                                            X    X
                                                                 X
X    X
                                      B-3
                                                                        OLM04.2

-------
Exhibit B--Section 1
Contract Reports/Deliverables Distribution
                                                            Distribution
            Item
No. of
Copies
                                     Delivery
                                     Schedule
                                                           o
                                                           w
                                                                o
                                    cn
                                    0)
            Standard
            Operating
            Procedures--
            Technical and
            Evidentiary
         .Revise within 60
         days after
         contract award.

         Submit within 7
         days of receipt
         of written
         request to
         recipients as
         directed.
                        -As directed
            Quality
            Assurance Plan
         Revise within 60
         days after
         contract award.
            GC/MS  and GC/EC
            Tapes
  Lot
Submit within 7
days of receipt
of written
request to
recipients as
directed.
Retain for 365
days after data
submission.
                                                             As directed
As directed
            Extracts
                               Lot
         Submit within 7
         days after
         receipt of
         written request
         by APO.
         Retain for 365
         days after data
         submission.
                        As directed
                                     Submit within 7
                                     days after
                                     receipt of
                                     written request
                                     by APO or SMO,
                                     at the Agency's
                                     direction.
                                      B-4
                                                                       OLM04.2

-------
                                                          Exhibit B--Section 1
                                    Contract Reports/Deliverables Distribution
Footnotes:
     AThe number of copies specified are the number of copies required to be
delivered to each recipient.

     BThe number of days associated with these elements will be provided in
the associated laboratory contract document, and will also be provided at the
time of the sample scheduling by the SMO' Contractor.

     cContractor-concurrent delivery to EPA designated recipient (e.g.,  QATS)
may be required upon request by the APO.  Retain for 365 days after data
submission, and submit as directed within 7 days after receipt of written
request by the APO.
     Also required in the Sample Data Summary Package.

     A sample delivery group  (SDG) is a group of samples within a Case,
     received over a period of  7 days or less and not exceeding 20 samples
      (excluding PE samples) .  Data, for all samples in the SDG are due
     concurrently.  The date  of delivery of the SDG or any samples within the
     SDG is the date that the last sample in the SDG is received.  See Exhibit
     A for further description.

     DELIVERABLES ARE TO BE REPORTED TOTAL AND COMPLETE.  Concurrent delivery
     required.  Delivery shall be made such that all designated recipients
     receive the item on the  same calendar day.  This includes resubmission of
     both the hardcopy and  electronic deliverable.  The date of delivery of
     the SDG, or any sample within the SDG, is the date all samples have been
     delivered.  If the deliverables are due on a Saturday, Sunday, or Federal
     holiday, then they shall be delivered on the next business day.
     Deliverables delivered after this time will be considered late.'

     Complete SDG File will contain the original•sample data package plus all
     of the original documents  described under Section .2.7.

     See Exhibit E and Exhibit  F for a more detailed description.

     If requested at the  time of sample scheduling, the Contractor shall
     provide Preliminary Results, consisting of Form I and Form I TIC
     analytical results, by fraction, for field and QC sample analyses via
      telefacsimile  (fax) 'or other electronic means.  The  Contractor will be
     notified of  the fax number or E-mail address at the  time of sample
     scheduling.  Sample  Traffic Reports and SDG cover sheets shall be
     submitted with  the Preliminary Results.  The Contractor shall contact SMO
     after  confirming  transmission.  The Contractor shall document all
      communication in  a  telephone contact log.
                                      B-5                               OLM04.2

-------
Exhibit B--Section 1
Contract Reports/Deliverables Distribution

Footnotes (con't):

     Preliminary Results Delivery Schedule:

     If the last sample in the SDG arrives before 5 p.m., the Preliminary
     Results are due within the required turnaround time.  If the last sample
     in the SDG is received after 5 p.m., the Preliminary Results are due
     within the required turnaround time beginning at 8 a.m. the following
     day.  DELIVERABLES ARE TO BE REPORTED TOTAL AND COMPLETE.  Concurrent
     delivery is required.  Delivery shall be made such that all designated
     recipients receive the item on the same calendar day.

     If the deliverables are due on a Saturday, Sunday, or Federal holiday,
     then they shall be delivered on the next business day.  Deliverables
     delivered after this time will be considered late.

NOTE:  As specified in the Contract Schedule (Government Furnished Supplies
and Materials), unless otherwise instructed by the CLP Sample Management
Office based on a Regional decision, the Contractor shall dispose of unused
sample volume and used sample bottles/containers no earlier than sixty (60)
days following submission of the reconciled Complete SDG File.  Sample
disposal and disposal of unused sample bottles/containers is the
responsibility of the Contractor, and should be done in accordance with all
applicable laws and regulations governing disposal of such materials.

1.2-  Distribution.  The following addresses correspond to the "Distribution"
      column in Table 1 of Section 1.1.

     SMO:        USEPA Contract Laboratory Program
                 Sample Management Office  (SMO)1
                 2000 Edmund Halley Drive
                 Reston, VA 20191-3436

     Region:     USEPA Region:  The Sample Management Office will provide the
                 Contractor with the list of addresses for the 10 EPA Regions.
                 SMO will provide the Contractor with updated Regional
                 address/name lists as necessary throughout the period of the
                 contract and identify other client recipients on a
                 case-by-case basis.

     QATS:       USEPA Contract Laboratory Program
                 Quality Assurance Technical Support  (QATS) Laboratory2
                 2700 Chandler Avenue, Building C
                 Las Vegas, NV  89120
                 Attn:  Data Audit Staff
      xThe Sample Management Office (SMO)  is a contractor operated facility
operating under the CLASS contract awarded and administered by the EPA.

      2The Quality Assurance Technical Support (QATS)  Laboratory is a
contractor operated facility operating under  the QATS contract awarded and
administered by the EPA.

                                      B-6                    .           OLM04.2

-------
                                                          Exhibit B--Section 2
                         Reporting Requirements and Order of Data Deliverables

2.0   REPORTING REQUIREMENTS AND ORDER OF DATA DELIVERABLES

2.1   Introduction.  The Contractor shall provide reports and other
      deliverables as specified in the Contract Schedule (Performance/Delivery
      Schedule).   The required content and form of each deliverable is
      described in this exhibit.  All reports and documentation must be:

      •    Legible,

      •   . Clearly labeled and completed in accordance with instructions in
          this exhibit,

      •    Arranged in the order specified in this section,

      •    Paginated consecutively in ascending order starting from the SDG
          Narrative, and

      •    Copies must be legible and double-sided.

      NOTE:  Complete SDG files need not be double-sided.   (The CSF is
      composed of original documents.)  However, sample data packages
      delivered to SMO must be double-sided.

2.1.1      Requirements for each deliverable item cited in the Contract
           Schedule  (Contract Performance/Delivery Schedule) are specified in
           Sections 2.3 through 2.11.  Prior to submission, the Contractor
           shall arrange items and the components of each item in the order
           listed in these sections.

2.1.2      The Contractor shall use EPA Case numbers  (including SDG numbers)
           and EPA sample numbers to identify  samples received under this
           contract, both verbally and in reports/correspondence.  The
           contract number shall be specified  in all correspondence.

2.2   Resubmission of Data.   If submitted documentation does not conform to
      the above criteria, the Contractor shall resubmit such documentation
      with deficiency(ies) corrected, at no additional cost to the Agency.

2.2.1      The Contractor shall respond within seven  (7) days to written
           requests  from data recipients for additional information or
           explanations that  result from the Government's inspection
           activities unless  otherwise specified in the contract.

2.2.2      Whenever  the Contractor is required to submit or resubmit data as  a
           result  of an on-site laboratory  evaluation, or through an
           Administrative Project Officer/Technical Project Officer action, or
           through a Regional data reviewer's  request, the data  shall be
           clearly marked as  ADDITIONAL DATA and shall be sent to both
           contractual data recipients  (SMO and the Region; and  to  the  EPA
           .designated  recipient  (e.g., QATS) when a written request for the
           sample  data package  has been made).  .The Contractor shall include  a
           cover  letter which describes which  data are being delivered, to
           which  EPA Case(s)  the  data pertain, and who requested the data.

 2.2.3      Whenever  the Contractor is required to submit or resubmit data as  a
           result  of Contract Compliance Screening  (CCS)  review  by  SMO, the

                                      B-7                               OLM04.2

-------
Exhibit B--Section 2
Reporting Requirements and Order of Data Deliverables

           data shall be sent to both contractual data recipients  (SMO and the
           Region; and to the EPA designated recipient (e.g., QATS) when a
           written request for the sample data package has been made).   In all
           instances, the Contractor shall include a color-coded COVER SHEET
           (Laboratory Response To Results of Contract Compliance Screening)
           provided by SMO.

2.3  Quality Assurance Plan and Standard Operating Procedures.  The Contractor
     shall adhere to the requirements in Exhibits E and F.

2.4  Sample Traffic Reports.  Each sample received by the Contractor will be
     labeled with an EPA sample number, and will be accompanied by a Sample
     Traffic Report  (TR) bearing the sample number and descriptive information
     regarding the sample.  The Contractor shall complete the TR  (marked "Lab
     Copy for Return to SMO"), recording the date of sample receipt and sample
     condition upon receipt for each container, and shall sign the TR.
     Information shall be recorded for each sample in the SDG.

2.4.1      The Contractor shall submit TRs in SDG sets (i.e., TRs for all
           samples in an SDG shall be clipped together),  with an SDG cover
           sheet attached.  The SDG cover sheet shall contain the following
           items:

           •      Laboratory name,

           •      Contract number,

           •      Sample analysis price  (full sample price from the contract),

           •      Case number, and

           •      List of EPA sample numbers of all samples in the SDG,
                 identifying the first and last samples received, and their
                 dates of receipt  (LRDs).

           NOTE:  When more than one sample is received in the first or last
           SDG shipment, the "first" sample received would be the lowest
           sample number  (considering both alpha and numeric designations);
           the "last" sample received would be the highest sample number
           (considering both alpha and numeric designations).

2.4.2      Each TR shall be clearly marked with the SDG number, entered below
           the laboratory receipt date on the TR.  The TR for the last sample
           received in the SDG shall be clearly marked "SDG--FINAL SAMPLE."
           The SDG number is the EPA sample number of the first sample
           received in the SDG.  When several samples are received together in
           the first SDG shipment, the SDG number shall be the lowest sample
           number  (considering both alpha and numeric designations) in the
           first group of samples received under the SDG.

2.4.3      If samples are received at the laboratory with multi-sample TRs,
           all the samples on one multi-sample TR may not necessarily be in
           the same SDG.  In this instance, the Contractor shall make the
           appropriate number of photocopies of the TR, and submit one copy
           with each SDG cover sheet.


                                      B-8                              OLM04.2

-------
                                                          Exhibit B--Section 2
                         Reporting Requirements.and Order of Data Deliverables


2.5  Sample Data Summary Package.  The sample data.summary package shall be
     ordered as follows and shall be submitted separately (i.e., separated by
     rubber bands, clips, or other means) directly preceding the sample data
     package.  Sample data forms shall be arranged in increasing EPA sample
     number order, considering both letters and numbers.  For example, BE400
     is a lower sample number than BF100, as E precedes F in the alphabet.
     The SDG number shall be reported on all data reporting forms.  The sample
     data summary package shall be arranged in the same manner as the sample
     data package.  The sample data summary package shall contain data for all
     samples in one SDG of the Case, as follows:   (See Section 2.6 for a
     detailed description of each item.)

2.5.1      SDG Narrative.

2.5.2      Arranged by fraction and by sample within each fraction:  tabulated
           target compound results  (Form I) for the volatile, semivolatile,
           and pesticide fractions and tentatively identified compounds  (Form
           I TIC) for the volatile and semivolatile fractions only.

2.5.3      Arranged by fraction:  system monitoring compound or  surrogate
           spike  analysis results  (Form II) by matrix  (water and/or soil)  for
           the volatile, semivolatile, and pesticide fractions;  and for soil,
           by concentration  (low or medium), for volatile and semivolatile
           fractions.

2.5.4      Arranged by fraction:  matrix spike/matrix spike duplicate results
           .(Form  III) for the volatile, semivolatile, and pesticide fractions.

2.5.5      Arranged by fraction:  blank data  (Form IV) and tabulated results
            (Form  I) for  the volatile, semivolatile, and pesticide  fractions
           including tentatively identified compounds  (Form I TIC)  for the
           volatile and  semivolatile  fractions only.

2.5.6      Arranged by fraction:   internal  standard area data  (Form VIII)  for
           the volatile  and  semivolatile fractions only.

2.6  Sample  Data  Package.  The  sample data package is divided into the five
     major units  described in this  section.  The last three units  are each
     specific to  an analytical  fraction  (volatiles, semivolatiles,  and
     pesticides/Aroclors).   If  the  analysis of a fraction is not required,
     then that fraction-specific unit is not required as a deliverable.   The
     sample  data  package shall  include data for the analyses of  all samples in
     one SDG, including:  field samples, dilutions, reanalyses,  blanks, matrix
     spikes, and  matrix  spike duplicates.   The Contractor shall  retain a  copy
     of the  sample data  package for 365  days after final acceptance of data.
     After this time, the Contractor may dispose of the package.

2.6.1    SDG Narrative.  This document  shall be clearly labeled "SDG
         Narrative"  and  shall contain:   laboratory name; Case number; EPA
         sample numbers  in the  SDG,  differentiating between  initial analyses
         and reanalyses; SDG number;  Contract  number,-  and detailed
         documentation of any quality control,  sample,  shipment, and/or
         analytical  problems encountered in processing  the  samples reported in
         the data package.   All volatile low level soil samples prepared


                                      B-9                               OLM04.2

-------
Exhibit B--Section 2
Reporting Requirements and Order of Data Deliverables

         according to the Modified SW-846 Method 5035 must be noted in the SDG
         Narrative.  When using the Modified SW-846 Method 5035, all
         discrepancies between sample weights determined in the field and in
         the laboratory shall be documented in the SDG Narrative.

         All GC columns used for analysis shall be documented here, by
         fraction.  List- the GC column identification--brand name, the
         internal diameter, in mm, and the length, in meters, packing/coating
         material and film thickness.  The trap used for volatile analysis
         shall be described here.  List trap name, when denoted by the
         manufacturer, its composition (packing material/brand name, amount of
         packing material, in length, cm).  All tentatively identified alkanes
         and their estimated concentrations are to be reported here.  The EPA
         sample number, the CAS number, when available, the alkane compound
          (or series) name, and its estimated concentration shall be provided
         in a tabular format.  The Contractor shall include any technical and
         administrative problems encountered, the corrective actions taken,
         the resolution, and an explanation for all flagged edits  (e.g.,
         manual edits) on quantitation lists.  The Contractor shall document
         in the SDG Narrative all instances of manual integration.

         The SDG Narrative shall contain the following statement, verbatim:
         "I certify that this data package is in compliance with the terms and
         conditions of the contract, both technically and for completeness,
         for other than the conditions detailed above.  Release of the data
         contained in this hardcopy data package and in the computer-readable
         data submitted on diskette has been authorized by the laboratory
         manager or his/her designee, as verified by the following signature."
         This statement shall be directly followed by an original signature'of
         the laboratory manager or his/her designee with a typed line below it
         containing the signer's name and title, and the date of signature.

2.6.1.1     Whenever data from sample reanalyses are submitted, the Contractor
            shall state in the SDG Narrative for each reanalysis whether the
            reanalysis is billable, and if so, why.

2.6.1.2     The Contractor shall list the pH determined for each water sample
            submitted for volatiles analysis.  This information may appear as
            a simple list or table in the SDG Narrative.  The purpose of this
            pH determination is to ensure that all water volatiles samples
            were acidified in the field.  No pH adjustment is to be performed
            by the Contractor on water samples for volatiles analysis.

2.6.2    Traffic Reports.  The Contractor shall include a copy of the TRs
         submitted in Section 2.4 for all of the samples in the SDG.  The TRs
         shall be arranged in increasing EPA sample number order, considering
         both letters and numbers.  Copies of the SDG cover sheet are to be
         included with the copies of the TRs.   (See Section 2.4 for more
         detail on reporting requirements for TRs.)  In the case of multi-
         sample TRs, the Contractor shall make the appropriate number of
         photocopies of the TR so that a copy is submitted with each
         applicable data package.  In addition, in any instance where samples
         from more than one multi-sample TR are in the same data package, the
         Contractor shall submit a copy of the SDG cover- sheet with copies of
         the TRs.


                                     B-10                              OLM04.2

-------
                                                          Exhibit B--Section 2
                         Reporting Requirements.and Order of Data Deliverables
                                                                     Volatiles

2.6.3    Volatiles Data

2.6.3.1     Volatiles QC Summary

2.6.3.1.1      System Monitoring Compound Summary  (Form II, VOA-1, VOA-2).

2.6.3.1.2      Matrix Spike/Matrix Spike Duplicate Summary  (Form III, VOA-1,
               VOA-2).

2.6.3.1.3      Method Blank Summary  (Form IV VOA):  If more than a single form
               is necessary, forms shall be arranged in chronological order by
               date of analysis of the blank, by  instrument.

2.6.3.1.4      GC/MS instrument performance check  (Form V VOA):  If more than
               a single form is necessary, forms  shall be arranged in
               chronological order, by instrument.

2.6.3.1.5      Internal Standard Area and RT Summary  (Form VIII VOA):  If more
               than a single form is necessary, forms shall be arranged in
               chronological order, by instrument.

2.6.3.2     Volatiles Sample Data.  Sample data shall be arranged in packets
            with the Organic Analysis Data Sheet  (Form I, VOA-1, VOA-2,
            including Form  I VOA-TIC), followed by the raw  data for volatile
            samples.  These sample packets shall  be placed  in order of
            increasing EPA  sample number order, considering both letters and
            numbers.

2.6.3.2.1      Target Compound Results, Organic Analysis Data Sheet  (Form I,
               VOA-1, VOA-2).  Tabulated results  (identification and
               quantitation) of the  specified target compounds  (Exhibit C,
               Volatiles) shall be included.  The validation and release of
               these results are authorized by a  specific,  signed statement in
               the  SDG Narrative  (see Section 2.6.1).  In  the event that the
               laboratory manager cannot verify all data reported for each
               sample, the  laboratory manager shall provide a detailed
               description  of the problems associated with  the sample in the
               SDG  Narrative.

2.6.3.2.2      Tentatively  Identified Compounds  (Form I VOA-TIC).   Form  I
               VOA-TIC is the tabulated list of the highest probable match  for
               up to  30 organic compounds that are not system monitoring
               compounds or internal standard compounds and are not  listed  in
               Exhibit C.   It includes the Chemical Abstracts  Service  (CAS)
               registry number  (if applicable), tentative  identification, and
               estimated concentration.  This form shall be included even if
               no compounds are found. If no compounds are  found, indicate
               this on the  form by entering  "0"  in the field  for  "Number
               Found."

2.6.3.2.3      Reconstructed Total Ion Chromatograms  (for  each sample or
               sample extract, including dilutions and reanalyses).
               Reconstructed ion  chromatograms shall be normalized  to the
                                      B-ll                              OLM04.2

-------
Exhibit B--Section 2
Reporting Requirements and Order of Data Deliverables
Volatiles

               largest nonsolvent component and shall contain the following
               header information:

               •   EPA sample number,

               •   Date and time of analysis,

               •   GC/MS instrument identifier,

               •   Lab file identifier, and

               •   Analyst ID.

2.6.3.2.3.1       Internal standards and system monitoring compounds shall be
                  labeled with the names of compounds, either directly out
                  from the peak or on a printout of retention times if
                  retention times are printed over the peak.

2.6.3.2.3.2       If automated data system procedures are used  for preliminary
                  identification and/or quantitation of  the  target compounds,
                  the complete data system report shall  be included in all
                  sample data packages, in addition to the reconstructed  ion
                  chromatogram.  The complete data system report shall include
                  all of the information listed below.   For  laboratories  which
                  do not use the automated data system procedures, a
                  laboratory "raw data sheet" containing the following
                  information shall be included in the sample data package, in
                  addition to the chromatogram:

                  •   EPA sample number,

                  •   Date and time of analysis,

                  •   Retention time or scan number of identified target
                      compounds,

                  •   Ion used for quantitation with measured area,

                  •   Copy of area table from data system,

                  •   GC/MS instrument identifier,

                  •   Lab file identifier, and

                  •   Analyst ID.

2.6.3.2.3.3       In all instances where the data system report has been
                  edited, or where manual integration or quantitation has been
                  performed, the GC/MS operator shall identify  such edits or
                  manual procedures by initialing and dating the changes  made
                  to the report, and shall include the integration scan range.
                  In addition, a hardcopy printout of the EICP  of the
                  quantitation ion displaying the manual integration shall be
                  included in the raw data.  This applies to all compounds


                                     B-12                              OLM04.2

-------
                                                          Exhibit B--Section 2
                         Reporting Requirements and Order of Data Deliverables
                                                                     Volatiles

                  listed in Exhibit C (Volatiles),:internal standards and
                  system monitoring compounds.

                  •   EICPs displaying each manual integration.

2.6.3.2.4      Other Required Information.  For each sample, by each compound
               identified, the following items shall be included in the data
               package.

               •   Copies of raw spectra and copies of background-subtracted
                   mass spectra of target compounds listed in Exhibit C
                    (Volatiles) that are identified in the sample and
                   corresponding background-subtracted target compound
                   standard mass spectra.  Spectra shall be labeled with EPA
                   sample number, lab file identifier, date and time of
                   analysis, and GC/MS instrument identifier.  Compound names
                   shall be clearly marked on all spectra.

               •   Copies of mass spectra of organic compounds not listed in
                   Exhibit C with associated best-match spectra  (minimum of
                   one, maximum of three best matches).  Spectra shall be
                   labeled with EPA sample number, lab file identifier, date
                   and time of analysis, and GC/MS instrument identifier.
                   Compound names shall be clearly marked on all spectra.

2.6.3.3     Volatiles Standards Data

2.6.3.3.1       Initial calibration data  (Form VI, VOA-1, VOA-2) shall be
                included in order by instrument,  if more than one instrument is
               used.

                •   Volatile standard(s) reconstructed  ion chromatograms and
                   quantitation reports for the  initial  (five-point)
                    calibration, labeled as in Section  2.6.3.2.3.  Spectra are
                   not required.

                •   All initial calibration data  that pertain to samples in the
                    data package shall be  included, regardless of when it was
                   performed  and  for which Case.   When more than one initial
                    calibration is performed,  the data  shall be in
                    chronological  order, by instrument.

                •    EICPs  displaying each  manual  integration.

2.6.3.3.2       Continuing calibration data  (Form VII,  VOA-1, VOA-2)  shall, be
                included in order  by instrument,  if more  than one instrument is
                used.

                •    Volatile standard(s)  reconstructed  ion  chromatograms and
                    quantitation  reports  for all  continuing  (12-hour)
                    calibrations,  labeled  as in Section 2.6.3.2.3.   Spectra are
                    not required.
                                      B-13                               OLM04.2

-------
Exhibit B--Section 2
Reporting Requirements and Order of Data Deliverables
Volatiles

               •   When more than one continuing calibration is performed,
                   forms shall be in chronological order, by instrument.

               •   EICPs displaying each manual integration.

2.6.3.3.3      In all instances where the data system report has been edited,
               or where manual integration or quantitation has been performed,
               the GC/MS operator shall identify such edits or manual
               procedures by initialing and dating the changes made to the
               report, and shall include the integration scan range.  In
               addition, a hardcopy printout of the EICP of the quantitation
               ion displaying the manual integration shall be included in the
               raw data.  This applies to all compounds listed in Exhibit C
                (Volatiles), internal standards and system monitoring
               compounds.

2.6.3.4     Volatiles Raw QC Data

2.6.3.4.1      BFB data shall be arranged in chronological order by instrument
               for each 12-hour period, for each GC/MS system utilized.

               •   Bar graph spectrum, labeled as in Section 2.6.3.2.3.

               •   Mass listing, labeled as in Section 2.6.3.2-. 3.

               •   Reconstructed total ion chromatogram, labeled as in Section
                   2.6.3.2.3.

2.6.3.4.2      Blank data shall be arranged by type of blank  (method, storage,
               instrument) and shall be in chronological order by instrument.

               NOTE:  This order is different from that used for samples.

                   Tabulated results  (Form I, VOA-1, VOA-2).

               •   Tentatively identified compounds  (Form I VOA-TIC) even if
                   none are found.

               •   Reconstructed ion chromatogram(s) and quantitation
                   report(s), labeled as in Section 2.6.3.2.3.

               •   Target compound spectra with laboratory-generated standard,
                   labeled as in Section 2.6.3.2.4.  Data systems which are
                   incapable of dual display shall provide spectra in the
                   following order:

                       Raw target compound spectra.

                       Enhanced or background-subtracted spectra.

                       Laboratory-generated standard spectra.
               •   GC/MS library search spectra for tentatively identified
                   compounds, labeled as in Section 2.6.3.2.4.
                                     B-14                              OLM04.2

-------
                                                          Exhibit B--Section 2
                         Reporting Requirements and Order of Data Deliverables
                                                                 Semivolatiles

                   Quantitation/calculation of tentatively identified compound
                   concentrations.
2.6.3.4.3      Volatiles Matrix Spike Data

                   Tabulated results (Form I, VOA-1, VOA-2) of target
                   compounds.  Form I VOA-TIC is not required.

               •   Reconstructed ion'chromatogram(s) and quantitation
                   report(s), labeled as in Section 2.6.3.2.3.  Spectra are
                   not required.

2.6.3.4.4      Volatiles Matrix Spike Duplicate Data

               •   Tabulated results (Form I, VOA-1, VOA-2) of target
                   compounds.  Form I VOA-TIC is not required.

               •   Reconstructed ion chromatogram(s) and quantitation
                   report(s), labeled as in Section 2.6.3.2.3.  Spectra are
                   not required.

2.6.4    Semivolatiles Data

2.6.4.1     Semivolatiles QC Summary

2.6.4.1.1      Surrogate Percent Recovery Summary  (Form II, SV-1, SV-2).

2.6.4.1.2      Matrix Spike/Matrix Spike Duplicate  Summary (Form  III,  SV-1,
               SV-2)

2.6.4.1.3      Method Blank  Summary  (Form IV SV):   If more than a single  form
               is necessary, forms shall be.arranged in chronological  order  by
               date of analysis of the blank, by  instrument.

2.6.4.1.4      GC/MS Instrument Performance Check  (Form V  SV):  If more than a
               single form is necessary, forms shall be arranged  in
               chronological order, by instrument.

2.6.4.1.5       Internal Standard Area and RT Summary  (Form VIII,  SV-1,  SV-2):
                If more than  a single form is necessary, forms  shall be
               arranged in chronological order, by  instrument.

2.6.4.2     Semivolatiles Sample Data.  Sample data shall  be  arranged  in
            packets with the Organic Analysis Data  Sheet  (Form I, SV-1, SV-2,
            including Form I SV-TIC), followed by the raw  data for
            semivolatile samples.  These sample packets shall  be  placed in
            increasing EPA sample number order,  considering both  letters  and
            numbers.

2.6.4.2.1       Target Compound  Results, Organic Analysis Data  Sheet  (Form I
                SV-1, SV-2).  Tabulated results  (identification and
                quantitation) of the  specified target compounds (Exhibit C,
                Semivolatiles) shall be .included.   The validation  and release
                of these results are  authorized by a specific,  signed statement


                                     B-15                               OLM04.2

-------
Exhibit B--Section 2
Reporting Requirements and Order of Data Deliverables
Semivolatiles

               in the SDG Narrative  (see Section. 2.6.1) .  In the event that
               the laboratory manager cannot verify all data reported for each
               sample, the laboratory manager shall provide a detailed
               description of the problems associated with the sample in the
               SDG Narrative.

2.6.4.2.2      Semivolatile Tentatively Identified Compounds (Form I SV-TIC).
               Form I SV-TIC is the tabulated list of the highest probable
               match for up to 30 of the non-surrogate/non-internal standard
               organic compounds that are not listed in Exhibit C  (Volatiles,
               Semivolatiles).  It includes the CAS registry number  (if
               applicable), tentative identification,  and estimated
               concentration.  This form shall be included even if no
               compounds are found.  If no compounds are found, indicate this
               on the form by entering "0" in the field for "number found."

2.6.4.2.3      Reconstructed Total Ion Chromatograms (for each sample,
               including dilutions and reanalyses).   Reconstructed ion
               chromatograms shall be normalized to the largest nonsolvent
               component and shall contain the following header information:

               •   EPA sample number,

               •   Date and time of analysis,

               •   GC/MS instrument identifier,

               •   Lab file identifier, and

               •   Analyst ID.

2.6.4'. 2.3.1       Internal standards and surrogate compounds shall be labeled
                  with the names of compounds, either directly out from the
                  peak or on a printout of retention times if retention times
                  are printed over the peak.

2.6.4.2.3.2       If automated data system procedures are used for preliminary
                  identification and/or quantitation of the target compounds,
                  the complete data system report shall be included in all
                  sample data packages, in addition to the reconstructed ion
                  chromatogram.  The complete data system report shall include
                  all of the information listed below.  For laboratories which
                  do not use the automated data system procedures, a
                  laboratory "raw data sheet" containing the following
                  information shall be included in the sample data package, in
                  addition to the chromatogram.

                  •   EPA sample number,

                  •   Date and time of analysis,

                  •   Retention time or scan number of identified target
                      compounds,
                                     B-16                              OLM04.2

-------
                                                          Exhibit B--Section 2
                         Reporting Requirements and Order of Data Deliverables
                                                                 Semivolatiles

                  •    Ion used for quantitation with.measured area,

                  •    Copy of area table from data system,

                  •    GC/MS instrument identifier, and

                  •    Lab file identifier.

2.6.4.2.3.3       In all instances where the data system report has been
                  edited, or where manual integration or quantitation has been
                  performed, the GC/MS operator shall identify such edits or
                  manual procedures by initialing and dating the changes made
                  to the report, and shall include the integration scan range.
                  In addition, a hardcopy printout of the EICP of the
                  quantitation ion displaying the manual integration shall be
                  included in the raw data.  This applies to all compounds
                  listed in Exhibit C (Semivolatiles), internal standards, and
                  system monitoring compounds.

                  •    EICPs displaying each manual integration.

2.6.4.2.4      Other Required Information.  For each sample, by each compound
               identified, the following shall be included in the data
               package.

               •   Copies of raw spectra and copies of background-subtracted
                   mass spectra of target compounds listed in Exhibit C
                    (Semivolatiles) that are identified in the sample and
                   corresponding background-subtracted target compound
                   standard mass spectra.  Spectra shall be labeled with EPA
                   sample number, lab file identifier, date and time of
                   analysis, and GC/MS instrument identifier compound names
                   shall be clearly marked on all spectra.

               •   Copies of mass spectra of non-surrogate/non-internal
                   standard organic compounds not listed  in Exhibit C
                    (Volatiles and Semivolatiles)  with associated best-match
                   spectra  (maximum of three best matches).  This includes the
                   mass spectra for tentatively identified alkanes.  Spectra
                   shall be labeled with EPA sample number, lab  file
                   identifier, date and time of analysis, and GC/MS instrument
                   identifier compound names shall be clearly marked on all
                   spectra.

2.'6.4. 3      Semivolatiles Standards Data

2.6.4.3.1       Initial  calibration data  (Form VI  SV-1, SV-2) shall  be included
                in order by  instrument, if more than one  instrument  used.

                •   Semivolatile standard(s) reconstructed ion chromatograms
                   and  quantitation reports for the initial  (five-point)
                   calibration, labeled as in  Section 2.6.4.2.3.  Spectra are
                   not  required.
                                     B-17                               OLM04.2

-------
Exhibit B--Section 2
Reporting Requirements and Order of Data Deliverables
Semivolatiles

               •   All initial calibration data that pertain to samples in the
                   data package shall be included, regardless of when it was
                   performed and for which Case.   When more than one initial
                   calibration is performed, the data shall be in
                   chronological order, by instrument.

               •   EICPs displaying each manual integration.

2.6.4.3.2      Continuing calibration data  (Form VII SV-1, SV-2) shall be
               included in order by instrument, if more than one instrument is
               used.

               •   Semivolatile standard(s) reconstructed ion chromatograms
                   and quantitation reports for all continuing  (12-hour)
                   calibrations, labeled as in Section 2.6.4.2.3.  Spectra are
                   not required.

               •   When more than one continuing calibration is performed,
                   forms shall be in chronological order, by instrument.

               •   EICPs displaying each manual integration-.

2.6.4.3.3      In all instances where the data system report has been edited,
               or where manual integration or quantitation has been performed,
               the GC/MS operator shall identify such edits or manual
               procedures by initialing and dating the changes made to the
               report, and shall include the integration scan range.  In
               addition, a hardcopy printout of the EICP of the quantitation
               ion displaying the manual integration shall be included in the
               raw data.  This applies to all compounds listed in Exhibit C
                (Semivolatiles), internal standards, and system monitoring
               compounds.

2.6.4.4     Semivolatiles Raw QC Data

2.6.4.4.1      DFTPP data shall be arranged in chronological order by
               instrument for each 12-hour period, for each GC/MS system
               utilized.

               •   Bar graph spectrum, labeled as in Section 2.6.4.2.3.

               •   Mass listing, labeled as in Section 2.6.4.2.3.

               •   Reconstructed total ion chromatogram, labeled as in Section
                   2.6.4.2.3.

2.6.4.4.2      Blank data shall be included in chronological order by
               extraction date.

               NOTE:  This order is different from that used for samples.

                   Tabulated results  (Form I SV-1, SV-2).
                                     B-18                              OLM04.2

-------
                                                           Exhibit B--Section 2
                          Reporting Requirements and Order of Data Deliverables
                                                                  Semivolatiles

                •   Tentatively  identified compounds  (Form I SV-TIC) even if
                    none  are  found.

                •   Reconstructed  ion  chromatogram(s) and quantitation
                    report(s), labeled as in Section 2.6.4.2.3.

                •   Target  compound spectra with laboratory-generated standard,
                    labeled as in  Section 2.6.4.2.4.  Data systems which are
                    incapable of dual  display  shall provide spectra  in  the
                    following order:

                        Raw target compound spectra.

                        Enhanced or background-subtracted spectra.

                        Laboratory-generated standard spectra.

                •   GC/MS library  search spectra for  tentatively identified
                    compounds, labeled as in Section  2.6.4.2.4.

                •   Quantitation/calculation of tentatively identified  compound
                    concentrations.

  2.6.4.4.3     Semivolatiles Matrix  Spike Data

                •  . Tabulated results  (Form  I  SV-1, SV-2) of  target  compounds.
                    Form  I SV-TIC  is  not required.

                •   Reconstructed  ion chromatogram(s) and quantitation
                    report(s),  labeled as in Section  2.6.4.2.3.   Spectra  are
                    not required.

  2.6.4.4.4      Semivolatiles Matrix  Spike Duplicate  Data

                 •   Tabulated results (Form  I  SV-1,  SV-2) of  target  compounds.
                    Form I SV-TIC  is  not required.

                 •   Reconstructed  ion chromatogram(s) and quantitation
                     report(s),  labeled as in Section  2.6.4.2.3.   Spectra  are
                    not required.

|  2.6.4.4.5      Semivolatile GPC Data.  The  two most  recent UV traces of the
j                 GPC calibration solution,  and the  reconstructed ion
                 chromatogram and data system reports  for the  GPC blank  shall  be
                 arranged in chronological order by GPC  for  the GPC calibration.

                 •    UV traces labeled with the GPC column identifier,  date of
                     calibration, and compound names.   Compound names shall be
                     placed directly out from the peak,  or on  the printout of
                     retention times when the retention  times  are printed
                     directly over the peak.
                                       B-19                              OLM04.2

-------
Exhibit B--Section 2
Reporting Requirements and Order of Data Deliverables
Pesticides/Aroclors

               •   Reconstructed ion chromatogram and data system report(s)
                   labeled as specified in Section 2.6.4.2.3 for GPC blank
                   analysis.

               •   Reconstructed ion chromatogram and data system report(s)
                   for all standards used to quantify compounds in the GPC
                   blank labeled as specified in Section 2.6.4.2.3  (continuing
                   calibration standard).

2.6.5    Pesticide/Aroclor Data

2.6.5.1     Pesticide/Aroclor QC Summary

2.6.5.1.1      Surrogate Percent Recovery Summary  (Form II, PEST-1, PEST-2).

2.6.5.1.2      Matrix Spike/Matrix Spike Duplicate Summary (Form III, PEST-1,
               PEST-2).

2.6.5.1.3      Method Blank Summary  (Form IV PEST):  If more than a single
               form is necessary, forms shall be arranged in chronological
               order by date of analysis of the blank.

2.6.5.2     Pesticide/Aroclor Sample Data.  Sample data shall be arranged in
            packets with the Organic Analysis Data Sheet  (Form I PEST), •
            followed by the raw data for pesticide samples.  These sample
            packets should then be placed in order of increasing' EPA sample
            number, considering both letters and numbers.

2.6.5.2.1      Target Compound Results, Organic Analysis Data Sheet  (Form I
               PEST).  Tabulated results  (identification and quantitation) of
               the specified target compounds  (Exhibit C, Pesticides/Aroclors)
               shall be included.  The validation and release of these results
               is authorized by a specific, signed statement in the SDG
               Narrative  (see Section 2.'6.1).  In the event that the
               laboratory manager cannot verify all data reported for each
               sample, the laboratory manager shall provide a detailed
               description of the problems associated with the sample in the
               SDG Narrative.

2.6.5.2.2      Copies of Pesticide Chromatograms.  Positively identified
               compounds shall be labeled with the names of compounds, either
               directly out from the peak on the chromatogram, or on a
               printout of retention times on the data system printout if
               retention times are printed over the peak on the chromatogram.
               All chromatograms shall meet the acceptance criteria in Exhibit
               D PEST, and shall be labeled with the following information:

               •   EPA sample number,

               •   Volume injected  (fj.Ii) ,

               •   Date and time of injection,
                                     B-20                              OLM04.2

-------
                                                          Exhibit B--Section 2
                         Reporting Requirements and Order of Data Deliverables
                                                           Pesticides/Aroclors

               •    GC column identifier (by stationary-phase and internal
                   diameter),

               •    GC instrument identifier, and

               •  f Scaling factor  (label the x and y axes using a numerical
                   scale).

2.6.5.2.3      Copies of pesticide chromatograms from the second GC column
               shall be included and labeled as in Section 2.6.5.2.2.

2.6.5.2.4      Data System Printout.  A printout of retention time and
               corresponding peak height or peak area shall accompany each
               chromatogram.   The printout shall be labeled with the EPA
               sample number.   In all instances where the data system report
               has been edited, or where manual integration or quantitation
               has been performed, the GC/EC operator must identify such edits
               or manual procedures by initialing and dating the changes made
               to the report,  and shall include the integration time range.

2.6.5.2.5      All manual work sheets shall be included in the sample data
               package.

2.6.5.2.6      Other Required Information.  If pesticides/Aroclors are
               confirmed by GC/MS, the Contractor shall submit copies of
               reconstructed ion chromatograms, raw spectra, and background-
               subtracted mass spectra of target compounds listed in Exhibit C
               (Pesticides/Aroclors) that are  identified in the sample and
               corresponding background-subtracted TCL standard mass spectra.
               Compound names shall be clearly marked on all spectra.  For
               multicomponent pesticides/Aroclors confirmed by GC/MS, the
               Contractor shall submit mass spectra of three major peaks of
               multicomponent compounds from samples and standards.

2.6.5.3     Pesticide/Aroclor Standards Data

2.6.5.3.1      Initial Calibration of Single Component Analytes  (Form VI PEST-
               1, PEST-2):  for all GC columns, all instruments, in
               chronological order by GC column and instrument.

2.6.5.3.2      Initial Calibration of Multicomponent Analytes  (Form VI PEST-
               3):   for  all GC columns, all instruments, in chronological
               order by  GC column  and instrument.

2.6.5.3.3      Analyte Resolution  Summary  (Form VI PEST-4):  for all GC
               columns and instruments, in  chronological, order by GC column
               and  instrument.

2.6.5.3.4      Performance Evaluation Mixture  (Form VI PEST-5):  for all GC
               columns and instruments, in  chronological order by GC column
               and  instrument.  .
                                      B-21                               OLM04.2

-------
Exhibit B--Section 2
Reporting Requirements and Order"of Data Deliverables
Pesticides/Aroclors

2.6.5.3.5      Individual-Standard Mixture A  (Form VI PEST-6):  for all GC
               columns and instruments, in chronological order by GC column
               and instrument.

2.6.5.3.6      Individual Standard Mixture B  (Form VI PEST-7):  for all GC
               columns and instruments, in chronological order by GC column
               and instrument.

2.6.5.3.7      Calibration Verification Summary  (Form VII PEST-1):  for all
               performance evaluation mixtures and instrument, blanks, on all
               GC columns and instruments, in chronological order by GC column
               and instrument.

2.6.5.3.8      Calibration Verification Summary  (Form VII PEST-2):  for all
               mid-point concentrations of Individual Standard Mixtures A and
               B and instrument blanks used for calibration verification, on
               all GC columns and instruments, in chronological order-by GC
               column and instrument.

2.6.5.3.9      Analytical Sequence  (Form VIII PEST):  for all GC columns and
               instruments, in chronological order by GC column and
               instrument.

2.6.5^-3.10     Florisil Cartridge Check  (Form IX PEST-1):  for all lots of
      -:-        cartridges used to process samples in the SDG.

2.6.5/3.11     Pesticide GPC Calibration Verification  (Form IX PEST-2):  for
               all GPC columns, in chronological order by calibration
               verification date.

2.6.5^3.12     Pesticide Identification Summary for Single Component Analytes
     Ti         (Form X PEST-1):  for all samples with positively identified
               single component analytes, in order by increasing EPA sample
               number.

2.6.5.3.13     Pesticide Identification Summary for Multicomponent Analytes
                (Form X PEST-2):  for all samples with positively identified
               multicomponent analytes, in order by increasing EPA sample
               number.

2.6.5.3.14     Chromatograms and data system printouts shall be included for
               all standards including the following:

               •   Resolution check mixture.

               •   Performance evaluation mixtures, all.

               •   Individual Standard Mixture A, at three concentrations,
                   each initial calibration.

               •   Individual Standard Mixture B, at three concentrations,
                   each initial calibration.
                                     B-22                              OLM04.2

-------
                                                          Exhibit B--Section 2
                         Reporting Requirements and Order of Data Deliverables
                                                           Pesticides/Aroclors

               •    All multicomponent analytes (toxaphene and Aroclors),  each
                   initial calibration.

               •    All mid-point concentrations of Individual Standard
                   Mixtures A and B used for calibration verification.

               •    All multicomponent analyte standards analyzed for
                   confirmation.

2.6.5.3.15     A printout of retention time and corresponding peak height or
               peak area shall accompany each chromatogram.  The printout
               shall be labeled with the EPA sample number.  In addition, all
               chromatograms shall meet the acceptance criteria in Exhibit D
               PEST, and shall be labeled with the following:

               •    EPA sample number for the standard.   (e.g., INDAL1, INDAM2,
                   etc.).  See Section 4 for details.

               •    Label all standard peaks for all individual compounds
                   either directly out from the peak on the chromatogram or on
                   the printout of retention times on the data system printout
                   if retention times are printed over the peak on the
                   chromatogram.

               •    Total nanograms injected for each standard.  When total
                   nanograms injected appear on the printout, it is not
                   necessary to include them on the chromatogram.

               •    Date  and time of injection.

               •    GC column identifier  (by stationary phase and internal
                   diameter).

               •    GC instrument identifier:

               •    Scaling factor  (label the x and y axes  using a numerical
                   scale).

               NOTE:  In all instances where the data system report has been
               edited, or where manual integration or quantitation has been
               performed, the GC/EC operator must identify such edits or
               manual procedures by initialing and dating  the changes made to
               the report, and  shall include the integration time range.

2.6.5.4     Pesticide/Aroclor Raw QC Data

2.6.5.4.1      Blank  data shall be arranged by type of blank  (method,
               instrument, sulfur cleanup) and shall be  in chronological order
               by  instrument.

               NOTE:  This order  is different from that  used for samples.

               •   Tabulated results  (Form I PEST).
                                     ; B-23                              OLM04.2

-------
Exhibit B--Section 2
Reporting Requirements and Order of Data Deliverabies
Pesticides/Aroclors

               •   Chromatogram(s)  and data system printout(s) for each GC
                   column and instrument used for analysis, labeled as in
                   Sections 2.6.5.2.2 and 2.6.5.2.4.

2.6.5.4.2      Matrix Spike Data

               •   Tabulated results  (Form I PEST) of target compounds.

               •   Chromatogram(s)  and data system printout(s), labeled as in
                   Sections 2.6.5.2.2 through 2.6.5.2.4.

2.6.5.4.3      Matrix Spike Duplicate Data

               •   Tabulated results  (Form I PEST) of target compounds.

               •   Chromatogram(s)  and data system printout(s), labeled as in
                   Sections 2.6.5.2.2 through 2.6.5.2.4.

2.6.5.5     Raw GPC Data

2.6.5.5.1      GPC Calibration.  The UV traces for the GPC calibration
               solution, chromatograms, and the data system reports for the
               GPC blank shall be arranged in chronological order for the GPC
               calibration.

               •   UV traces labeled with the GPC column identifier, date of
                   calibration, and compound names.  Compound names shall be
                   placed directly out from the peak, or on the printout of
                   retention times when the retention times are printed '
                   directly over the peak.

               •   Chromatogram and data system report(s) labeled as specified
                   in Sections 2.6.5.2.2 and 2.6.5.2.4 for GPC blank analyses.

               •   Chromatogram and data system report(s) for all standards
                   used to quantify compounds in the GPC blank labeled as
                   specified in Section 2.6.5.3.15  (i.e., Individual Standard
                   Mixture A, Individual Standard Mixture B, and the
                   Aroclor/toxaphene standards).

2.6.5.5.2      GPC Calibration Verification.  The Chromatogram and the data
               system report(s) shall be arranged in chronological order for
               the GPC calibration check.

               •   Chromatograms and data system printouts  labeled as
                   specified in Sections 2.6.5.2.2 and 2.6.5.2.4 for the GPC
                   calibration verification solution analyses.

               •   Chromatogram and data system report(s) for standards used
                   to quantify compounds in the GPC calibration verification
                   solution or used to assess the Aroclor pattern labeled as
                   specified in Section 2.6.5.3.15  (i.e., Individual Standard
                   Mixtures A and B and Aroclor Standard Mixture 1016/1260
                   from the initial calibration sequence).


                                     B-24                              OLM04.2

-------
                                                          Exhibit B--Section 2
                         Reporting Requirements and Order of Data Deliverables

2.6.5.6     Raw Florisil Data.  The chromatogram and data system report(s)
            shall be arranged in chronological order by Florisil cartridge
            performance check analyses.

            •    Chromatograms and data system reports labeled as specified in
                Sections 2.6.5.2.2 and 2.6.5.2.4 for the florisil cartridge
                performance check analyses.

            •    Chromatograms and data system reports for standard analyses
                used to quantify compounds in the Florisil cartridge
                performance check analysis,  labeled as specified in Section
                2.6.5.3.15  (i.e., Individual Standard Mixture A and Individual
                Standard Mixture B and the 2,4,5 Trichlorophenol solution).

2.7   Complete SDG File.  As specified in Section 1, the Contractor shall
      deliver one Complete SDG File  (CSF) including the original sample data
      package to the Region concurrently with delivery of the sample data
      package to SMO.  Delivery to EPA designated recipient  (e.g., QATS) is
      only required upon written request.

2.7.1    The CSF will contain all original documents specified in Sections 3
         and 4 and in Form DC-2  (see Section 4).  No photocopies of original
         documents will be placed in the CSF unless the original data was
         initially written in a bound notebook, maintained by the Contractor,
         or the originals were previously submitted to the Agency with another
         Case/SDG in accordance with the requirements described in Exhibit F.
         The contents of the CSF shall be numbered according to the
         specifications described in Section 3.20.

2.7.2    The CSF will consist of the following  original documents in addition
         to the documents in the sample  data package.

         NOTE:  All SDG-related documentation may be used or admitted as
         evidence in subsequent legal proceedings.  Any other SDG-specific
         documents generated after the CSF is sent to EPA, as well as copies
         that are altered in any fashion, are also deliverables to EPA.
         Deliver the original to the Region and a copy to SMO.  Delivery  to
         EPA designated recipient  (e.g., QATS)  is only .upon written  request.

2.7.2.1     The original sample data package.

2.7.2.2     A completed and signed document inventory sheet  (Form DC-2).

2.7.2.3     All original shipping documents including, but not limited to, the
            following documents:

            •   EPA Chain-of-Custody Record,

            •   Airbills  (if  an  airbill  is not  received,  include  a hardcopy
                receipt requested from the shipping company or a printout of
                 the shipping  company's electronic  tracking  information),

            •   EPA Traffic Reports, and

            •   Sample  tags  (if present) sealed in plastic bags.


                                     ,B-25                              OLM04.2

-------
Exhibit B--Section 2
Reporting Requirements and Order of Data Deliverables
2.7.2.4     All original receiving documents including, but not limited to,
            the following documents:

            •   Form DC-1,

            •   Other receiving forms or copies of receiving logbooks, and

            •   SDG cover sheet.

2.7.2.5     All original laboratory records, not already submitted in the
            sample data package,  of sample transfer, preparation, and analysis
            including, but not limited to, the following documents:

            •   Original preparation and analysis forms or copies of
                preparation and analysis logbook pages,

            •   Internal sample and sample extract transfer chain-of-custody
                records,

            •   Screening records, and

            •   All instrument output, including strip charts from screening
                activities.

2.7.2.6     All other original SDG-specific documents in the possession of the
            Contractor including, but not limited to, the following documents:

            •   Telephone contact logs,

            •   Copies of personal logbook pages,

            •   All handwritten SDG-specific notes, and

            •   Any other SDG-specific documents not covered by the above.

2.7.3    If the Contractor does submit SDG-specific documents to EPA after
         submission of the CSF, the documents should be identified with unique
         accountable numbers, a revised Form DC-2 should be submitted, and the
         unique accountable numbers and locations of the documents in the CSF
         should be recorded in the "Other Records" section on the revised Form
         DC-2.  Alternatively, the Contractor may number the newly submitted
         SDG-specific documents to EPA as a new CSF and submit a new Form DC-
         2.  The revised Form DC-2 or new Form DC-2 should be submitted to the
         EPA Regions only.

2.8   Data in Computer-Readable Format.  The Contractor shall provide a
      computer-readable copy of the data on data reporting Forms I-X for all
      samples in the SDG as specified in Exhibit H, and delivered as specified
      in the Contract Schedule (Contract Performance/Delivery Schedule).
      Computer-readable data deliverables shall be submitted on IBM or
      IBM-compatible, 3.5-inch high-density 1.44 M-byte diskette  (or via an
      alternate means of electronic transmission approved in advance by the
      EPA) .
                                     B-26                              OLM04.2

-------
                                                          Exhibit B--Section 2
                         Reporting Requirements and Order of Data Deliverables

2.8.1    When submitted, the diskette(s) shall be packaged and shipped in such
         a manner that the diskette(s)  cannot be bent or folded, and will not
         be exposed to extreme heat or cold or any type of electromagnetic
         radiation.  The diskette(s)  shall be included in the same shipment as
         the hardcopy data and shall, at a minimum, be enclosed in a diskette
         mailer.  The diskette(s) shall be labeled as specified in Exhibit H,
         Section 8.4.

2.8.2    The data shall be recorded in ASCII, text file format, and shall
         adhere to the file, record,  and field specifications listed in
         Exhibit H.

2.9   Preliminary Results.  The Form I data results shall be submitted for all
      samples in one SDG of a Case.  This includes tabulated target compound
      results  (Form I) for the volatile, semivolatile, and pesticide
      fractions, and tentatively  identified compounds  (Form I TIC) for the
      volatile and semivolatile fractions.  The Contractor shall clearly
      identify the Preliminary Results by labeling each Form I  and Form I TIC
      as  "Preliminary Results" under each form title  (e.g., under Volatile
      Organics Analysis Data Sheet, Volatile Organics Analysis  Data Sheet
      Tentatively Identified Compounds).

2.10  GC/MS and GC/EC Tapes.  The Contractor shall adhere to the requirements
      in Exhibit E.

2.11  Extracts.  The Contractor shall preserve sample extracts  at 4° C (± 2°C)
      in bottles/vials with Teflon-lined septa.  Extract bottles/vials shall
      be labeled with EPA sample  number. Case number, and SDG number.  The
      Contractor shall maintain a logbook of stored extracts, listing EPA
      sample numbers and associated Case and SDG numbers.  The  Contractor
      shall retain extracts for 365 days following submission of the
      reconciled complete sample  data package.  During that time, the
      Contractor shall  submit extracts  and associated logbook pages within
      seven days following receipt of a written request  from the
      Administrative Project Officer or Technical  Project Officer.
                                      B-27                              OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
General Information

3.0   FORMS INSTRUCTIONS

3.1   Introduction.  This"section includes specific instructions for
      completing the data reporting forms required under this contract.  Each
      of the forms are specific to a given fraction (volatile, semivolatile,
      or pesticide/Aroclor) and, in some instances, specific to a given matrix
      (water or soil) within each fraction.  The Contractor shall submit only
      those forms pertaining to the fractions analyzed for a given sample(s).
      For instance, if a sample is scheduled for volatiles analysis only, the
      Contractor shall provide only forms for the volatile fraction.  NOTE:
      There are two pages relating to the volatile fraction for Forms I, VI,
      and VII.  There are also two pages relating to the semivolatile fraction
      for Forms I, VI, VII, and VIII.  Whenever volatiles or semivolatiles are
      analyzed and one of these forms is required, both pages  (VOA-1 and VOA-
      2; SV-1 and SV-2) shall be submitted.

3.2   General Information.  The Contractor shall report values on the hardcopy
      forms according to the individual form instructions in this section.
      For example, results for concentrations of volatile target compounds
      shall be reported to two significant figures if the value is greater
      than or equal to 10.  Values that exceed the maximum length allowed
      shall be reported to the maximum possible, maintaining the specified
      decimal place.

3.2.1    The data reporting forms presented in Section 4 have been designed in
         conjunction with the computer-readable data format specified in
         Exhibit H.  The specific length of each variable for
         computer-readable data transmission purposes is also given in Exhibit
         H.  Information entered on these forms shall not exceed the size of
         the field given on the form, including such laboratory-generated
         items as lab name and lab sample identifier.

         NOTE:  The space provided for entries on the hardcopy forms (Section
         4) is greater in some instances than the length prescribed for the
         variable as written to the electronic deliverable  (see Exhibit H).
         Greater space is provided on the hardcopy forms for visual clarity.

3.2.2    When submitting data, the Contractor shall reproduce all characters
         that appear on the data reporting forms in Section 4.  The format of
         the forms submitted shall be identical to that shown in the contract.
         No information may be added, deleted, or moved from its specified
         position without prior written approval of the Administrative Project
         Officer.  The names of the various fields and compounds (e.g., "Lab
         Code," "Chloromethane") shall appear as they do on the forms in the
         contract, including the options specified in the form (e.g., "Matrix:
          (soil/water)" shall appear, not just "Matrix").  For items appearing
         on the uncompleted forms  (Section 4), the use of uppercase and
         lowercase letters is optional.

3.2.3    Alphabetical entries made on the forms by the Contractor shall be in
         ALL UPPERCASE letters  (e.g., "LOW", not "Low" or "low").  If an entry
         does not fill the entire blank space provided on the form, null
         characters shall be used to remove the remaining underscores that
         comprise the blank line.  See Exhibit H for more detailed


                                     B-28                              OLM04.2

-------
                                                          Exhibit B--Section 3
                                                           . Forms Instructions
                                                           General Information

         instructions.  However,  the Contractor shall not remove the
         underscores or vertical bars that delineate "boxes" on the forms.
         The only exception would be those underscores at the bottom of a
         "box" that are intended as a data entry line.  (For instance,  on Form
         2A, line 30, if data is entered on line 30, it will replace  the
         underscores.)

3.3   Header Information.  Six pieces of information are common to the header
      section of each data reporting form:  lab name, contract, lab code, case
      number, SAS number, and SDG number.  Except as noted for SAS number,
      this information shall be entered on every form and shall match on every
      form.

3.3.1    Lab Name.  The lab name shall be the name chosen by the Contractor to
         identify the laboratory.  It shall not exceed 25 characters.

3.3.2    Contract.  Contract refers to the number of the EPA contract under
         which the analyses were performed.

3.3.3    Lab Code.  The lab code is an alphabetical abbreviation of up to six
         letters, as assigned by EPA, to identify the laboratory and aid in
         data processing.  This lab code will be assigned by EPA at .the time a
         contract is awarded, and shall not be modified by the Contractor,
         except at the direction of EPA.  If a change of name or ownership
         occurs at the laboratory, the lab code will remain the same until the
         Contractor  is directed by EPA to use another lab code.

3.3.4    Case Number.  The Case number is the EPA-assigned Case number
         associated  with the sample.  This number is reported on the Traffic
         Report.

3.3.5    SAS Number.  The SAS number is the EPA-assigned number for analyses
         performed under Special Analytical Services.  If samples are td be
         analyzed under SAS only and reported on these forms, then enter the
         SAS number'and leave the Case number blank.  If samples are analyzed
         according to the Routine Analytical Services  (RAS) protocols and have
         additional  SAS requirements, list both the  Case number and the  SAS
         number on all forms.  If there are no SAS  requirements, leave the
         "SAS No." field blank.

         NOTE:  Some samples in an SDG may have a SAS number while others may
         not.

3.3.6    SDG Number.  The  "SDG No." field is for the  sample delivery group
         number.   It is  the EPA sample number of a  field sample assigned to
         the  SDG  and shall be unique for each SDG within a  Case.  If fractions
         of the same field samples are scheduled under different  turnaround
         times,  thus creating separate SDGs containing the  same sample
         numbers,  a  different sample number shall be utilized  in  the
         assignment  of  the SDG number for each sample delivery group.  If a
         situation arises  where  there are an insufficient number  of samples
         for assignment  of SDG numbers  (i.e., 1 sample with a  7 day  turnaround
         for volatile analyses and a 14  day turnaround for  semivolatile  and
                                      B-29                              OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
General Information

         pesticide/Aroclor analyses), the Contractor shall contact SMO for the
         assignment of an SDG number.

3.3.7    Sample Number.  This number appears either in the header information
         of the form, or as the left column of a table summarizing data from a
         number of samples.  When the EPA sample number is entered in the
         triple-spaced box in the upper righthand corner of Form I, Form IV,
         or Form X, it should be entered on the middle line of the three lines
        .that comprise the box.

3.3.7.1     The Contractor shall identify all samples, including:  dilutions,
            reanalyses, matrix spikes, matrix spike duplicates, blanks, and
            standards with an EPA sample number.  For field samples, matrix
            spikes and matrix spike duplicates, the EPA sample number is the
            unique identifying number given in the Traffic Report that
            accompanied that sample.  In order to facilitate data assessment,
            the Contractor shall use the following sample suffixes:

            XXXXX      =  EPA sample number
            XXXXXMS    =  Matrix spike sample
            XXXXXMSD   =  Matrix spike duplicate sample
            XXXXXRE    =  Re-extracted and reanalyzed sample
            XXXXXDL    =  The suffix DL is appended to the EPA
                          sample number to indicate that  the
                          analytical results are a result of a
                          dilution of the original analysis             ,
                          (reported as EPA sample XXXXX).   See
                          Exhibit D for requirements for
                          dilutions.

3.3.7.2     There may be instances  when all samples analyzed must be listed on
            the form, regardless of whether or not they are part of the SDG
            being reported  (e.g., Form VIII PEST).  In these instances, use
            ZZZZZ as the EPA sample number for any sample analysis not
            associated with the SDG being reported.

3.3.7.3     For blanks, the Contractor shall use the following identification
            scheme for the EPA sample number:

            •   Volatile method blanks shall be identified as VBLK##.

            •   Volatile instrument blanks shall be identified as VIBLK##.

            •   Volatile storage blanks shall be identified as VHBLK##.

            •   Semivolatile-method blanks shall be identified as SBLK##.

            •   Pesticide/Aroclor method blanks and/or sulfur cleanup blanks
                shall be identified as PBLK##.

            •   Pesticide/Aroclor instrument blanks shall be identified as
                PIBLKtftf.
                                     B-30                              OLM04.2

-------
3.3.7.4
3.3.7.5
3.3.7.6
                                              Exhibit B--Section 3
                                                Forms Instructions
                                               General Information

The EPA sample number shall be unique for each blank within an
SDG.  Within a fraction, the Contractor shall achieve this by
replacing the two-character terminator (##) of the identifier with
one or two characters or numbers, or a combination of both.  For
example,, possible identifiers for volatile blanks would be VBLK1,
VBLK2, VBLKA1, VBLKB2,  VBLK10, VBLKAB, etc.  If the method blank
is analyzed on multiple instruments, then an additional two-
character suffix shall be added to make the blank EPA sample
number unique.

Volatile and semivolatile standards shall be identified as
FSTD***##, where
               STD  =
               ***  _
            Fraction code  (V for volatiles; S for semivolatiles).

            Standard.

            Concentration of volatile standards in ug/L  (e.g.,
            010, 020, 050, 100, and 200) or the amount injected  in
            ng for semivolatile standards  (e.g., 020, 050, 080,
            120, and 160).

            One,or two characters, numbers, or combinations of
            both to create a unique EPA sample number within an
            SDG.
The Contractor shall use the following scheme to identify
pesticide/Aroclor standards:
                            Name
                                   EPA Sample Number
              Individual Mix A (low point)
              Individual Mix A (mid-point)
              Individual Mix A (high point)
              Individual Mix B (low point)
              Individual Mix B (mid-point)
              Individual Mix B (high point)
              Resolution Check
              Performance Evaluation Mixture
              Toxaphene
              Aroclor 1016
              Aroclor 1221
              Aroclor 1232
              Aroclor 1242
              Aroclor 1248
              Aroclor 1254
              Aroclor 1260"
              Aroclor 1016/1260 Mix
                                        INDAL##
                                        INDAM##
                                        INDAH##
                                        INDBL##
                                        INDBM##
                                        INDBH##
                                         RESC##
                                         PEMtt#
                                        TOXAPH##
                                        AR1016##
                                        AR1221##
                                        AR1232##
                                        AR1242##
                                        AR1248##
                                        AR1254##
                                        AR1260##
                                        AR1660##
                                      B-31
                                                            OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
General Information

            The Contractor shall replace the two-character terminator (##) of
            the identifier with one or two characters or numbers, or a
            combination of both, to create a unique EPA sample number within
            an SDG.

3.3.7.7     If the standards are injected onto both GC columns on the same
            instrument simultaneously, the same EPA sample number may be used
            for reporting data for the standards for both columns.  If
            simultaneous injections are not made, then the same number shall
            not be used.

3.3.7.8     The EPA sample number for GPC shall be GPC#########, where
            ######### is the GPC column ID.  If the GPC column ID is more than
            nine characters, truncate at the ninth character.

3.3.7.9     The EPA sample number for florisil shall be FLO#######tt#, where
            ######### is the florisil cartridge lot number.  If the florisil
            cartridge lot number is more than nine characters, truncate at the
            ninth character.

3.3.8    Other Common Fields.  Several other pieces of information are common
         to many of the data reporting forms.  These include matrix, sample
         weight/volume, level, lab sample identifier, and lab file identifier.

         •   In the "Matrix" field, enter SOIL for soil/sediment samples and
             WATER for water samples.

             NOTE:  The matrix shall be spelled out.  Abbreviations such as S
             or W shall not be used.

         •   In the "Sample wt/vol" field, enter the number of grams  (for
             soil) or milliliters  (for water) of sample used in the first
             blank.  Enter the units, either G or ML, in the second blank.

         •   The "Level" field is used for the volatile and semivolatile
             fractions.  Enter the determination of concentration level made
             from the screening of soils.  Enter as LOW or MED, not L or M.
             All water samples shall be entered as LOW.

             NOTE:  There is no differentiation between low and medium soil
             samples for the pesticide/Aroclor fraction, and no level is
             entered on any of these forms.

         •   The lab sample identifier is a unique laboratory-generated
             internal identifier pertaining to a particular analysis.  The
             Contractor can enter up to 12 alpha-numeric characters in the
             "Lab Sample ID" field.  The Contractor may use the EPA sample
             number as the lab sample identifier.

         •   The lab file identifier is the unique laboratory-generated name
             of the GC/MS data system file containing information pertaining
             to a particular analysis.  The Contractor can enter up to 14
             alpha-numeric characters in the "Lab File ID" field.
                                     B-32                              OLM04.2

-------
                                                          Exhibit B--Section 3
                                                            Forms Instructions
                                                                        Form I

3.3.8.1     The "Instrument ID" field is common to the forms containing
            calibration data.  The identifier used by the Contractor shall
            include some indication of the manufacturer and/or model of the
            instrument, and shall contain additional characters that
            differentiate between all instruments of the same type in the
            laboratory.

3.3.8.2     Forms II, IV, V, VIII, IX, and X contain a field labeled "page _
            of _" in the bottom lefthand corner.  If the number of entries
            required on any of these forms exceeds the available space,
            continue entries on another copy of the same fraction-specific
            form, duplicating all header information.  If a second page is
            required, number the pages consecutively  (i.e., "page 1 of 2" and
            "page 2 of 2").  If a second page is not required, number the page
            "page 1 of 1."

            NOTE:  These  forms are fraction-specific, and often matrix-
            specific within a fraction.  For example, Form II VOA-1 and Form
            II VOA-2 are  for different data.  Therefore, do not number the
            pages of all  six versions of Form II as "1 of 6," "2 of 6," etc.
            Number only pages corresponding to  the fraction-specific and
            matrix-specific form.

3.3.9    Rounding Rule.   For rounding off numbers to the appropriate level of
         precision, the Contractor shall follow these rules.  If the figure
         following those  to be retained is less than 5, drop it  (round down).
         If the  figure is greater than or equal to 5, drop it and increase the
         last digit to be retained by 1  (round  up).

3.4    Organic Analysis Data Sheet  (Form I, All  Fractions)

3.4.1    Purpose.  This  form is used  for tabulating and reporting sample
         analysis, including blank, matrix spike, and matrix spike  duplicate
         results  for target compounds.  If all  fractions are not requested for
         analysis, only  the pages for the fractions required shall  be
         submitted.  For  example, if  only volatiles analysis is  requested,
         Form I,  VOA-1, VOA-2  and Form I VOA-TIC shall be  submitted.   If  only
         the pesticide/Aroclor fraction is requested  for analysis,  Form I PEST
         shall be submitted.   Furthermore, pesticide  instrument  blanks
          (PIBLKs)  shall be reported on a per column/per analysis basis on Form
         I  PEST.   Each PIBLK shall be named with a unique  EPA sample number.

3.4.2    Instructions..  Complete  the  header information according to the
         instructions  in Section  3.3.  Complete the remainder of the form
         using the following instructions.

3.4.2.1      For  soil samples analyzed-for volatiles,  enter the non-decanted
             percent  moisture in  the  "% Moisture: not  dec." field on Form  I,
             VOA-1, VOA-2.  This  is the only percent moisture determination
             made for volatiles since  the entire contents of the  VOA vial  are
             considered as the  sample.  For water samples,  leave  this field
             blank.
                                      B-33                              OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form I

3.4.2.2     For soil samples analyzed for semivolatiles and
            pesticides/Aroclors, enter the values for the percent moisture
            determined during the analysis in the "% Moisture" field on Form I
            SV-1, SV-2, or PEST.  In the "decanted  (Y/N)" field, enter Y if
            the sample had standing water above the soil/sediment that was
            decanted, or N if no water was decanted off the surface of the
            sample.  Report percent moisture  (decanted or not decanted) to the
            nearest whole percentage point (e.g., 5%, not 5.3%).  For water
            samples, method blanks, sulfur cleanup blanks, and instrument
            blanks, leave these fields on Form I blank.

3.4.2.3     For volatiles, enter the GC column identifier in the "GC Column"
            field on Form I, VOA-1, VOA-2, and the internal diameter in
            millimeters  (mm), to two decimal places, in the "ID" field.  For
            packed columns, convert the internal diameter from inches to
            millimeters as necessary before entering in the "ID" field.

3.4.2.4     For semivolatiles and pesticides/Aroclors, enter the method of
            extraction in the "Extraction" field on Form I SV-1, SV-2, SV-TIC,
            and PEST as SEPF for separatory funnel, CONT for continuous
            liquid-liquid extraction without hydrophobia membrane, CONH for
            continuous liquid-liquid extraction with hydrophobic membrane,
            SONG for sonication  (soils only), SOXH for Automated Soxhlet
            Extraction  (soils only), or PFEX for Pressurized Fluid Extraction
             (soils only) .

3.4.2.5     If gel permeation chromatography  (GPC) was performed, enter Y in
            the "GPC Cleanup" field on Form I SV-1, SV-2, or PEST.  Enter N in
            this field if GPC was not performed.

            NOTE:  GPC is required for all soil samples analyzed for
            semivolatiles and pesticides/Aroclors; therefore, all forms for
            soil samples will contain a Y in this field.

3.4.2.6     For soil samples only, enter the pH for semivolatiles and
            pesticides/Aroclors, reported to 0.1 pH units, on Form I SV-1,
            SV-2, or PEST.

3.4.2.7     Enter the date of sample receipt at the laboratory, as noted on
            the Traffic Report  (i.e., the VTSR), in the "Date Received" field.
            The date shall be entered as MM/DD/YY.

3.4.2.8     Complete the "Date Extracted" and "Date Analyzed" fields in the
            same format  (MM/DD/YY).  When continuous liquid-liquid extraction
            procedures are used for water samples, enter the date that the
            procedure was started in the "Date Extracted" field.  If
            separatory funnel  (pesticides only), sonication, soxhlet, or
            pressurized  fluid procedures are used, enter the date that the
            procedure was completed in the "Date Extracted" field.  For
            pesticide/Aroclor samples, enter the date of the first GC analysis
            performed in the "Date Analyzed" field.  The date of sample
            receipt will be compared with the extraction and analysis dates of
            each fraction to ensure that contract holding times were not
            exceeded.


                                     B-34                              OLM04.2

-------
                                                          Exhibit B--Section 3
                                                            Forms Instructions
                                                                        Form I

3.4.2.9     If a medium soil sample is analyzed for volatiles, enter total
            volume of the methanol extract in microliters  (uL) in the "Soil
            Extract Volume" field on Form I, VOA-1, VOA-2.  This volume
            includes any methanol not collected from the filtration of the
            extract through glass wool; the volume is typically 10,000 uL
            (i.e., the 10 mL of methanol used for the extraction).  If a
            medium soil sample is analyzed,, enter the volume of the methanol
            extract added to the reagent water in the purge tube and analyzed
            in the "Soil Aliquot Volume" field.  Enter this volume in
            microliters (uL).
                                                                   9
3.4.2.10    For semivolatiles and pesticides/Aroclors, enter the actual volume
            of the most concentrated sample extract, in microliters  (uL), in
            the "Concentrated Extract Volume" field on Form I SV-1, SV-2 or
            PEST.  For semivolatiles, this volume will typically be 1,000 uL
            (for water) or 500 uL  (for water and soil) when GPC is performed.
            For pesticides/Aroclors, the volume of the most concentrated
            extract will typically be 10,000 uL  (for water) or 5,000 uL  (for
            water and soil) when GPC is performed.  For pesticides/Aroclors,
            the volume of the most concentrated extract is not the volume
            taken through the Florisil and sulfur cleanup  steps.  If a
            dilution of the sample extract is made in a subsequent analysis,
            this volume will remain the same, but the dilution factor will
            change.

3.4.2.11    For semivolatiles and pesticides/Aroclors, enter  the volume of the
            sample extract  injected into the GC in the "Injection Volume"
            field on Form I SV-1, SV-2 or PEST.  Report this  volume in
            microliters  (uL) to one decimal place  (e.g.,  1.0  uL).

            NOTE:  A 2.0 microliter injection  is required for semivolatile
            analyses.

3.4.2.12    If pesticides/Aroclors are analyzed using two GC  columns connected
            to a  single injection port, enter  the amount  of half the volume  in
            the syringe in  the  "Injection Volume"  field  (i.e., assume that the
            extract injected is evenly divided between the two columns).

3.4.2.13    If a  sample or  sample extract has  been diluted for analysis, enter
            the dilution factor as a single number  (e.g.,  enter  100.0 for a  1
            to 100 dilution of  the sample)  in  the  "Dilution Factor"  field.
            The dilution factor shall not be entered  as a fraction.  If a
            sample was not  diluted, enter 1.0.  Report dilution  factors to one
            decimal place.

3.4.2.14    If sulfur  cleanup is employed,  enter Y in the "Sulfur Cleanup"
            field; if  not,  enter N on  Form  I PEST.

3.4.2.15    For positively  identified  target compounds, the Contractor  shall
            report the concentrations  as uncorrected  for  blank contaminants.

3.4.2.16    For volatile and semivolatile results, report analytical results
            to one significant  figure  if the value is less than  10,  and two
                                      B-35                              OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form I

            significant figures if the value is 10 or above.  Report all
            pesticide/Aroclor results to two significant figures.

3.4.2.17    Enter the appropriate concentration units, ug/L or ug/Kg.

3.4.2.18    Under the column labeled "Q" for qualifier, flag each result with
            the specific data reporting qualifiers listed below.  When
            reporting results to EPA, the Contractor shall use these contract-
           'specific qualifiers.  The Contractor shall not modify the
            qualifiers.  Up to five qualifiers may be reported on Form I for
            each compound.  The Contractor is encouraged to use additional
            flags or footnotes  (see the X qualifier).

            The EPA-defined qualifiers to be used are:

             U:  This flag indicates the compound was analyzed for but not
                 •detected.  The CRQL shall be adjusted according to the
                 equation listed in Exhibit D.  CRQLs are listed in Exhibit C.

             J:  This flag indicates an estimated value.  This flag is used
                  (1) when estimating a concentration for tentatively
                 identified compounds where a 1:1 response is assumed, (2)
                 when the mass spectral and retention time data indicate the
                 presence of a compound that meets the volatile and
                 semivolatile GC/MS identification criteria, and the result is
                 less than the CRQL but greater than zero, and  (3) when the
                 retention time data indicate the presence of a compound that
                 meets the pesticide/Aroclor identification criteria, and the
                 result is less than the CRQL but greater than zero.  For
                 example, if the sample quantitation limit is 10 ug/L, but a
                 concentration of 3 ug/L is calculated, report it as 3J.

                 NOTE:  The J flag is not used and the compound is not
                 reported as being identified for pesticide/Aroclor results
                 less than the CRQL if the pesticide residue analysis expert
                 determines that the peaks used for compound identification
                 resulted from instrument noise or other interferences (column
                 bleed, solvent contamination, etc.).

             N:  This flag indicates presumptive evidence of a compound.   This
                 flag is only used for tentatively identified compounds
                  (TICs), where the identification is based on a mass spectral
                 library search.  It is applied to all TIC results.  For
                 generic characterization of a TIC, such as chlorinated
                 hydrocarbon, the N flag is not used.

             P:  This flag is used for a pesticide/Aroclor target analyte when
                 there is greater than 25% difference for detected
                 concentrations between the two GC columns  (see Form X).   The
                 lower of the two values is reported on Form I and flagged
                 with a P.

             C:  This flag applies* to pesticide results where the
                 identification has been confirmed by GC/MS.  If GC/MS


                                     B-36                              OLM04.2

-------
                                            Exhibit B--Section 3
                                              Forms Instructions
                                                           Form I

    confirmation was  attempted but was unsuccessful, do not apply
    this  flag;  use  a  laboratory-defined  flag instead  (see  the  X
    qualifier).

B:  This  flag is used when the analyte is  found  in the  associated
    method blank as well  as in the sample.  It indicates probable
    blank contamination and warns the data user  to take
    appropriate action.   This flag shall be used for a
    tentatively identified compound  as well as for a positively
    identified target compound.

    The combination of flags BU or UB is expressly prohibited.
    Blank contaminants are flagged B only  when they are detected
    in the sample.

E:  This flag identifies  compounds whose concentrations exceed
    the upper level of the calibration range of  the instrument
    for that specific analysis.  If  one  or more  compounds  have a
    response greater  than the upper  level  of the calibration
    range, the sample or  extract shall be  diluted and  reanalyzed.
    according to the  specifications  in Exhibit D; exceptions are
    also noted in Exhibit D.  All such compounds with  a response
    greater than the  upper level of  the  calibration range  shall
    have the concentration flagged with  an E on  Form  I  for the
    original analysis.

    NOTE:  For total  xylene, where three isomers are  quantified
    as two peaks, the calibration range  of each  peak  shall be
    considered separately.  For example, a diluted analysis is
    not required for  total xylene unless the concentration of the
    peak representing the single isomer  exceeds  200 ug/L or the
    peak representing the two co-eluting isomers on that GC
    column exceeds 400 ug/L.

D:  If a sample or extract is reanalyzed at a  higher  dilution
    factor, for example when the concentration of an analyte
    exceeds the upper calibration range, the DL  suffix is
    appended to the sample number on Form I for  the more diluted
    sample, and all reported concentrations on that Form I are
    flagged with the  D flag.  This flag  alerts data users that
    any discrepancies between the reported concentrations may be
    due to dilution of the sample or extract.   NOTE 1:   The D
    flag is not applied to compounds which are not detected in
    the sample analysis  (i.e., compounds reported with the CRQL
    and the U flag).   NOTE 2: Separate Form Is are required for
    reporting the original analysis (EPA Sample No. XXXXX) and
    the more diluted sample analysis (EPA Sample No.  XXXXXDL)
    (i.e., the results from both analyses cannot be combined on a
    single Form I).

A:  This flag indicates that a tentatively identified.compound is
    a-suspected aldol-condensation product.
                        B-37                              OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form I

             X:  Other specific flags may be required to properly define the
                 results.'  If used, the flags shall be fully described, with
                 the description attached to the sample data summary package
                 and the SDG Narrative.  Begin by using X.  If more than one
                 flag is required, use Y and Z as needed.  If more than five
                 qualifiers are required for a sample result, use the X flag
                 to represent a combination of several flags.  For instance,
                 the X flag might combine the A, B, and D flags for some
                 samples.  The laboratory-defined flags are limited to X, Y,
                 and Z.

3.5   Organic Analysis Data Sheet:  Tentatively Identified Compounds  (Form I
      VOA-TIC and Form I SV-TIC)

3.5.1    Purpose.  This form is used to report analysis results for non-target
         compounds  (e.g., compounds not listed in Exhibit C), excluding system
         monitoring compounds, surrogates, and internal standards.  See
         Exhibit D for instructions on identification and quantitation.  The
         Contractor shall submit Form I VOA-TIC or SV-TIC for every analysis,
         including required dilutions and reanalyses, even if no TICs are
         found.

3.5.2    Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Complete the remainder of the form
         using the following instructions in addition to the instructions in
         Section 3.4.

3.5.2.1     Report all TICs including CAS number (if applicable), compound
            name, retention time, and the estimated concentration as
            uncorrected for blank contaminants.  If the analytical result is
            less than 10, report to one significant figure.  If the analytical
            result is 10 or greater, report to two significant figures.-
            (Criteria for reporting TICs are given in Exhibit D, Section 11).
            Retention time shall be reported in minutes and decimal minutes,
            not seconds or minutes:seconds.  If, in the opinion of the mass
            spectral interpretation specialist, no valid tentative
            identification can be made, the compound shall be reported as
            unknown.

3.5.2.2     Total the number of TICs found, including aldol-condensation
            products  (see Section 3.5.2.4), and enter this number in the
            "Number TICs found" field.  If no TICs were found, enter 0 (zero).

3.5.2.3     If the name of a compound exceeds the 28 spaces in the TIC column,
            truncate the name to 28 characters.  If the compound is an
            unknown, restrict the description to ho more than 28 characters
            (e.g., unknown hydrocarbon).

3.5.2.4     Peaks that are suspected to be aldol-condensation reaction
            products  (e.g., 4-methyl-4-hydroxy-2-pentanone and
            4-methyl-3-pentene-2-one) shall be summarized on this form,
            flagged A, and included in the "Number TICs found" field.  .The


                                     B-38                              OLM04.2

-------
                                                          Exhibit B--Section 3
                                                            Forms Instructions
                                                                       Form II

            peaks shall be counted as part of the 30 most intense non-target
            semivolatile compounds to be searched. .

3.6   System Monitoring Compound Recovery (Form II, VOA-1, VOA-2)

3.6.1    Purpose.  For volatiles, Form II, VOA-1, VOA-2 is used to report the
         recoveries of the system monitoring compounds added to each volatile
         sample, including dilutions and reanalyses, blank, matrix spike, and
         matrix spike duplicate.  The system monitoring compounds are used to
         monitor the performance of the purge and trap-gas chromatograph-mass
         spectrometer system as a whole.  Form II VOA is matrix-specific, so
         that system monitoring compound recoveries for water samples are
         reported on a different version of Form II than the recoveries for
         soil samples.  Soil sample recoveries are further differentiated by
         concentration level.

3.6.2    Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Complete one form for each level.  Do
         not mix low and medium level samples on one form.  NOTE:  For
         volatile soil samples only, specify the level as LOW or MED.
         Complete the remainder of the form using the following instructions.

3.6.2.1     For each system monitoring compound listed in Table 1, report the
            percent recovery to the nearest whole percentage point, and to the
            number of significant figures given by the QC limits at the bottom
            of the form.

3.6.2.2     Flag each system monitoring compound recovery outside the QC
            limits with an asterisk  (*).  The asterisk shall be placed in the
            last space in each appropriate column, under  the "#" symbol.

3.6.2.3     In the "TOT OUT" column, total the number of  system monitoring
            compound recoveries  that were outside  the QC  limits  for each
            sample.  If no system monitoring compounds were outside the
            limits, enter 0  (zero).

3.6.2.4     Number all pages as  described in Section 3.3.

                                    Table 1
                          System Monitoring Compounds
             Volatile System                    CAS Number
             Monitoring Compounds	


             SMC 1: Toluene-d8 (TOL)             2037-26-5
             SMC 2: Bromofluorobenzene (BFB)     460-00-4
             SMC 3: l,2-Dichloroethane-d4 (DCE)  17060-07-0	

 3.7    Surrogate Recovery (Form II,  SV-1,  SV-2  and Form II,  PEST-1,  PEST-2)

 3.7.1    Purpose.   Form II,  SV-1,  SV-2 and Form II,  PEST-1,  PEST-2 are used to
          report the recoveries of  the surrogate compounds added to each

                                      B-39                              OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form II

         semivolatile and pesticide/Aroclor sample, blank, matrix spike, and
         matrix spike duplicate.  Form II SV and Form II PEST are
         matrix-specific as well as fraction-specific, so surrogate recoveries
         for semivolatile and pesticide water samples are reported on a
         different version of Form II than surrogate recoveries for
         semivolatile and pesticide soil samples.

3.7.2    Instructions.  Complete the header information according to the
         instructions in Section 3.3.  NOTE:  For semivolatile soil samples
         only, specify the level as LOW or MED.  Complete one form for each
         level.  Do not mix low and medium level samples on one form.
         Complete the remainder of the form using the following instructions.

3.7.2.1        For each surrogate listed in Tables 2 and 3, report the percent
               recovery to the nearest whole percentage point.

3.7.2.2        Flag each surrogate recovery outside the QC limits with an
               asterisk (*).   The asterisk shall be placed in the last space
               in each appropriate column, under the "#" symbol.

3.7.2.3        In the "TOT OUT" column, total the number of surrogate
               recoveries that were outside the QC limits for each sample.  If
               no surrogates were outside the limits, enter 0 (zero).

3.7.2.4        If the sample is diluted and the surrogates are outside the
               acceptance window in any analysis, .enter the calculated
               recovery, and flag the surrogate recoveries with a D in the
               column under the "#" symbol.  Do not include results flagged
               with a D in the total number of recoveries for each sample
               outside the QC limits.

3.7.2.5        The pesticide surrogate recoveries shall be reported from both
               GC columns used for the analyses.  Therefore, identify each GC
               column at the top of Form II, PEST-1, PEST-2, entering the
               stationary phase in the "GC Column" field, and the internal
               diameter of the column in millimeters  (mm) in the "ID" field.

3.7.2.6        The assignment of columns as "1" and "2" is left to the
               discretion of the Contractor when the analyses are performed by
               simultaneous injection into a GC containing two columns.  If so
               analyzed, the assignment of "GC Column 1" and "GC Column 2"
               shall be consistent across all the reporting forms.  If the
               analysis is not performed by simultaneous injection, then the
               assignment of GC column number shall be based on the
               chronological order of the two analyses.

3.7.2.7        Although the pesticide surrogate recovery limits for samples,
               matrix spike and matrix spike duplicates are only advisory, the
               Contractor shall flag those recoveries that are outside the
               advisory QC limits or are diluted out.  The total number of
               recoveries that are outside the QC limits shall include all
               values from both of GC columns.  In counting the total number
               of recoveries that are outside the QC limits, do not include
               the results flagged with a D.


                                     B^40                              OLM04.2

-------
                                                          Exhibit B--Section 3
                                                            Forms Instructions
                                                                      Form III
3.7.2.8        Number all pages as described in Section 3.3.

                                    Table  2
                            Semivolatile Surrogates
          Semivolatile  Surrogates
                                 CAS Number
          SI
          S2
          S3
          S4
          S5
          S6
          S7
          S8
Nitrobenzene-d5 (NBZ)            4165-60-0
2-Fluorobiphenyl  (FBP)           321-60-8
Terphenyl-dl4  (TPH)              98904-43-9
Phenol-d5  (PHL)                  4165-62-2
2-Fluorophenol  (2FP)             367-12-4
2,4,6-Tribromophenol  (TBP)       118-79-6
2-Chlorophenol-d4  (2CP)          93951-73-6
l,2-Dichlorobenzene-d4  (DCS)	2199-69-1
                                    Table 3
                             Pesticide Surrogates
        Pesticide Surrogates
                                CAS Number
        Decachlorobiphenyl (DCB)
        Tetrachloro-m-xylene (TCX)
                                2051-24-3
                                877-09-8
3.8.  Matrix Spike/Matrix  Spike Duplicate Recovery  (Form  III, All  Fractions)

3.8.1    Purpose.  This  form  is used  to  report the  results  of the  analyses of
         matrix spikes and matrix  spike  duplicates  (MS/MSD).  The  form  is
         matrix-specific for  volatiles,  semivolatiles,  and  pesticides.

3.8.2    Instructions.   Complete the  header  information according  to the
         instructions in Section 3.3.  Include the  EPA  sample number for the
         matrix spike, without the suffixes  MS or MSD.  Complete the remainder
         of the form using the following instructions.

3.8.2.1     For volatile and  Semivolatile soil samples, specify level as LOW
            or MED on Form III, VOA-2, and SV-2.  SDGs  containing  soil  samples
            at both  levels require a  MS/MSD  at each level;  therefore, for
            soils, prepare one form for  each level.

3.8.2.2     In the first table under  the "SPIKE ADDED"  column,  enter the
            calculated concentration  in  ug/L or ug/Kg (according to the
            matrix)  that results from dividing each spike compound amount
            added to the aliquot weight/volume chosen for the matrix spike.
            For  instance,  for base/neutral compounds in medium  level soils,  if
            50 ug of spike are added  to  1 g  of soil,  the  resulting
            concentration  is  50,000 ug/Kg.
                                      B-41
                                                                        OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form IV

3.8.2.3     Enter the sample concentration in the next column, in similar
            units, of each spike compound detected in the original sample.  If
            a spike compound was not detected during the analysis of the
            original sample, enter the sample result as 0  (zero).

3.8.2.4     In the "MS CONCENTRATION" column, enter the actual concentration
            of each spike compound detected in the matrix spike aliquot.

3.8.2.5     Calculate the percent recovery of each spike compound in the
            matrix spike aliquot to the nearest whole percent, according to
            Exhibit D.  Enter the percent recovery in the "MS % REC" column.

3.8.2.6     Flag all percent recoveries outside the QC limits with an asterisk
            (*).  The asterisk shall be placed in the last space of the "MS %
            REC" column, under the "#" symbol.

3.8.2.7     For pesticide/Aroclor matrix spikes and matrix spike duplicates,
            the MS concentration and MSD concentration shall be the
            concentration of the spiked analyte reported on Form I for those
            analyses.  Of the two concentrations calculated for each
            pesticide/Aroclor target compound, one on each GC column, the
            lower concentration shall be reported on Form I, and both
            concentrations shall be reported on Form X..  The lower
            concentration is also reported on Form III and used in the
            calculation of spike recovery, even if that concentration yields a
            recovery value that is outside the advisory QC limits.

3.8.2.8     Follow Sections 3.8.2.2 through 3.8.2.7 to complete the lower
            table, using the results of the analysis of the MSD aliquot.

3.8.2.9     Calculate the relative percent difference (RPD) between the matrix
           . spike recovery and the matrix spike duplicate recovery, and enter
            this value in the "% RPD" column.  Report the RPD to the nearest
            whole percent.

3.8.2.10    Compare the RPDs to the QC limits given on the form, and flag each
            RPD outside the QC limits with ah asterisk (*) in the last space
            of the "% RPD" column, under the "#" symbol.

3.8.2.11    Summarize the values outside the QC limits at the bottom of the
            page.  No further action is required by the Contractor.

3.9   Method Blank Summary  (Form IV, All Fractions)

3.9.1     Purpose.  This form summarizes the samples associated with each
          method blank analysis.  The Contractor shall submit the appropriate
          Form  IV for each blank.

3.9.2     Instructions.  Complete the header information according to the
          instructions in Section 3.3.  The EPA sample number entered in the
          upper righthand corner shall be the same number entered on Form I for
          the blank.  Complete the remainder of the form using the following
          instructions.
                                     B-42                              OLM04.2

-------
                                                          Exhibit B--Section 3
                                                            Forms Instructions
                                                                       Form IV

3.9.2.1     Complete the following fields:  "Instrument ,10, " "Date Analyzed,"
            and "Time Analyzed."  Dates shall be entered as MM/DD/YY.  The
            time shall be reported in military.time.

3.9.2.2     Pesticide/Aroclor .contaminants shall meet the identification
            criteria requiring analysis of the blank on two different GC
            columns  (see Exhibit D PEST).   Enter the date, time, and
            instrument ID of both analyses of the blank on the pesticide
            method blank summary  (Form IV PEST).  The information on the two
            analyses is differentiated as Date Analyzed  (1), Date Analyzed
            (2), etc.  If the analyses were run simultaneously, the order of
            reporting is not important, but shall be consistent with the
            information reported on all other pesticide forms.  Otherwise,
            Date Analyzed  (1) shall indicate the analysis on column 1, and
            Date Analyzed  (2) shall indicate the analysis on column 2.

3.9.2.3     Identify the GC column and internal diameter  in the appropriate
            fields.

3.9.2.4     For volatiles, indicate the purging method by entering Y for
            heated purge or N for ambient temperature purge in  the  "Heated
            Purge:  Y/N" field on Form IV VGA.

3.9.2.5     For pesticide/Aroclor blanks, enter the method of extraction as
            SEPF for separatory funnel, CONH for continuous liquid-liquid
            extraction with hydrophobia membrane, CONT  for continuous
            liquid-liquid  extraction  without hydrophobic membrane,  SONC for
            sonication, SOXH for  automated soxhlet extraction or PFEX for
            pressurized fluid extraction on Form IV PEST.

3.9.2.6     For semivolatile and pesticide/Aroclor method blanks, enter the
            date of  extraction of the blank on Form IV  SV or PEST.

3.9.2.7     If the samples associated with pesticide/Aroclor blank  are
            subjected to sulfur cleanup,  then the blank shall also  be
            subjected to sulfur cleanup.  If sulfur cleanup is  employed, enter
            Y in the  "Sulfur Cleanup" field; if not, enter N on Form IV PEST.
            If only  some of the samples associated with the method  blank are
            subjected to sulfur cleanup,  a separate sulfur  cleanup  blank is
            required (see  Exhibit D PEST).  If a separate sulfur cleanup blank
            is prepared, complete one version of Form  IV associating all the
            samples  with the method blank, and a second version of  Form IV
            listing  only those  samples  associated with the  separate sulfur
            cleanup  blank.  NOTE:  Subjecting all samples associated with a
            method blank to  sulfur cleanup avoids the  need  for  two  forms.

3.9.2.8     For all  three  fractions,  as appropriate, summarize  the  samples
            including storage and volatile instrument  blanks,  associated with
            a given  method blank  in the table,  entering the EPA sample number
            and lab  sample identifier.  For volatiles,  enter  the lab file
            identifier  and the  time of  analysis of each sample.  For
            semivolatiles,  enter  lab  file identifier and the  date of analysis.
            For pesticides/Aroclors,  enter the  dates of both  analyses as Date
            Analyzed (1) and Date Analyzed  (2), as discussed  previously.


                                      B-43                             OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form V
3.9.2.9     For pesticide/Aroclor fraction, enter the lab file identifier only
            if GC/MS confirmation was attempted.  Otherwise, leave this field
            blank.

3.9.2.10    Number all pages as described in Section 3.3.

3.10  GC/MS:Instrument Performance Check and Mass Calibration (Form V VGA and
      Form V SV)

3.10.1   Purpose.  This form is used to report the results of the GC/MS
         instrument performance check for the volatile and semivolatile
         fractions and to summarize the date and time of analyses of samples,
         including dilutions, reanalyses, standards, blanks, matrix spikes,
         and matrix spike duplicates associated with each analysis of the
         instrument performance check solution.

3.10.-2   Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Complete the remainder of the form
         using the following instructions.

3.10.2.1    Enter the date and time of injection of the instrument performance
            check solution  (BFB for volatiles--CAS Number 460004, DFTPP for
            semivolatiles--CAS Number 5074715).  The date shall be entered as
            MM/DD/YY.  The time shall be reported as military time.

3.10.2.2    For volatiles, identify the GC column and internal diameter on
            Form V VGA.

3.10.2.3    For each ion listed on the form, enter the percent relative
            abundance in the righthand column of the first table.  Report
            relative abundances to the number of significant figures given for
            each ion in the ion abundance criteria column.

            NOTE:  For both BFB and DFTPP,  one or more of the high mass ions
            may exceed the abundance of the ion listed on the form as the
            nominal base peak, m/z 95 for BFB and m/z 198 for DFTPP.  Despite
            this possibility, all ion abundances shall be normalized to the
            nominal base peaks listed on Form V (see Exhibits D and E).

3.10.2.4    All relative abundances shall be reported as a number.  If the
            relative abundance is zero, enter 0, not a dash or other
            non-numeric character.  Where parentheses appear, compute the
            percentage of the ion abundance of the mass given in the
            appropriate footnote, and enter that value in the parentheses.

3.10.2.5    In the lower table, list all samples,  including dilutions and
            reanalyses, standards, blanks,  matrix spikes, and matrix spike
            duplicates analyzed under that instrument performance check in
            chronological order, by time of analysis (in military time).
            Refer to Section 3.3.7 for specific instructions for identifying
            standards and blanks.
                                     B-44                             . OLM04.2

-------
                                                          Exhibit B--Section 3
                                                            Forms Instructions
                                                                       Form VI

3.10.2.6    Complete the following fields for all standards, samples,
            including dilutions and reanalyses, blanks, matrix spikes, and
            matrix spike duplicates:  "EPA Sample No.," "Lab Sample ID," "Lab
            File ID," "Date Analyzed," and "Time Analyzed."

3.10.2.7    Number all pages as described in Section 3.3.

3.11  GC/MS Initial Calibration Data  (Form VI, VOA-1, VOA-2 and Form VI, SV-1,
      SV-2)

3.11.1   Purpose.  After a GC/MS system has undergone an initial five-point3
         calibration at the specific concentration levels described in Exhibit
         D, and after all initial calibration criteria have been met,  the
         Contractor shall complete and submit this form for each volatile or
         semivolatile target compound initial calibration performed which is
         relevant to the samples, including dilutions and reanalyses,  blanks,
         matrix spikes, or matrix spike duplicates in the SDG, regardless of
         when that calibration was performed.

3.11.2   Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Enter the Case number and SDG number
         for the current data package, regardless of the original Case for
         which the initial calibration was performed.  Complete the remainder
         of the form using the following instructions.

3.11.2.1    Enter the date(s) of the calibration.  If the calendar date
          ..- changes, during the calibration procedure, the inclusive dates
            shall be recorded.  Dates shall be entered as MM/DD/YY.

3.11.2.2    Enter the injection times of the first and last of the standards
            analyzed in the  "Calibration Times"  field.  Times shall be
            reported in military time.

3.11.2.3    For volatiles, complete the  "GC Column"  and  "ID" fields.  Indicate
            the purging method by entering  "Y" for heated purge or  "N"  for
            ambient  temperature purge in the  "Heated Purge:  (Y/N)" field.

3.11.2.4    Enter the lab file identifier for  each of the five calibration
            standards injected.  Complete the  response factor data for  the
            five calibration points,  and then  calculate and report the  average
            relative response factor  (RRF)  for all target compounds.

3.11.2.5    For volatiles, report the relative response  factors for the  system
            monitoring compounds in the  calibration  standards.,  For
            semivolatiles, report the response factors for  all surrogate
      3For semivolatiles, eight compounds  (2,4-Dinitrophenol, 2,4,5-
 Trichlorophenol,  2-Nitroaniline,  3-Nitroaniline,  4-Nitroaniline,
 4-Nitrophenol,  4,6-Dinitro-2-Methylphenol,  and  Pentachlorophenol)  will  only
 require  a  four-point  initial  calibration at 50,  80,  120,  and  160  total
 nanograms  because detection at  less  than 50 ng  per  injection  is difficult.   If
 a  four-point  calibration is performed for these compounds,  leave  the  "RRF20"
 column blank.

                                      B-45                              OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form VI

            compounds in the calibration standards.  The Contractor shall
            report the relative standard deviation (%RSD) for all compounds.
            See Exhibit D for equations.

3.12  GC/EC Initial Calibration Data (Form VI, PEST-1, PEST-2)

3.12.1   Purpose.  The initial calibration of pesticides/Aroclors involves the
         determination of retention times, retention time windows, and
         calibration factors.  For single component pesticide target
         compounds, these data are calculated from the analyses of the
         Individual Standard Mixtures A and B at three different concentration
         levels.  For the multicomponent target compounds, these data are
         calculated from a single point calibration.

3.12.2   Instructions.  Complete one Form VI for each GC column used for the
         three analyses of Individual Standard Mixture A  (low point, mid-
         point, and high point) and the three analyses of Individual Standard
         Mixture B during an initial calibration.  Complete the header
         information according to the instructions in Section 3.3.  Complete
         the remainder of the form using the following instructions.

3.12.2.1    In the "Level (x low)" field, enter the concentration of the .low
            point, mid-point, and high point calibration standards as a
            multiplier of the low point.  Therefore, for the low point, enter
            "1.0."  The concentration of the mid-point standard is specified
            in Exhibit D as four times the low point; therefore, enter "4.0."
            The high point standard shall be at least 16 times the low point,
            but may be higher, if that value lies within the linear range of
            the instrument,  as specified in Exhibit D-.  Therefore, enter the
            appropriate multiplier to the high point standard concentration to
            one decimal place.

3.12.2.2    Identify the GC column and internal diameter  (in millimeters, mm)
            in the appropriate fields.

3.12.2.3    Enter the dates of analysis of the first and last of the six
            standards on each form in the "Date(s) Analyzed" field.  Dates
            shall be entered as MM/DD/YY.

3.12.2.4    For each standard analyzed, enter the retention time of each
            applicable analyte in minutes and decimal minutes, under the
            appropriate concentration level in the "RT OF STANDARDS" column on
            Form VI PEST-1.

3.12.2.5    Calculate the mean retention time of each analyte from the three
            individual mixtures, and report it in the "MEAN RT" column on Form
            VI PEST-1.

3.12.2.6    Calculate the retention time window for each analyte using the
            specifications in Exhibit D, and enter the lower limit of the
            window in the "RT WINDOW" column under "FROM," and the upper limit
            of the window under "TO" on Form VI PEST-1.'  The retention times
            of the surrogates are reported from the analyses of Individual
                                     B-46                              OLM04.2

-------
                                                          Exhibit B--Section 3
                                                            Forms Instructions
                                                                       Eorm VI

            Mixture A and the windows are only required to be calculated for
            Individual Mixture A.    .               ;  • •  .

3.12.2.7    For the six analyses of the Individual Standard Mixtures, the
            Contractor shall also complete the calibration factor data on Form
            VI PEST-2.  Prepare one form for each instrument and GC column
            used.  Enter the calibration factor for each compound in each of
            the standards.   Calculate and enter a mean calibration factor and
            a relative standard deviation (%RSD).  As with surrogate retention
            times, the surrogate calibration factors are only required from
            Individual Mixture A analyses.  .

3.12.2.8    For the multicomponent target compounds, the retention times,
            .retention time windows, and calibration factors shall be reported
            in a similar fashion for each single point calibration standard.
            For each multicomponent compound, the Contractor shall select at
            least three peaks from each analyte, according to the
            specifications in Exhibit D.  The retention time and calibration
            factor data apply to each peak.  Complete one version of Form VI
            PEST-3 for each GC column,  for each initial calibration that
            applies to samples in the data package.

3.12.3   Form VI is also used to report the results of analysis of the
         Resolution Check Standard that shall begin each pesticide/Aroclor
         initial calibration sequence  (Form VI PEST-4).  The Contractor shall
         submit one Form VI PEST-4 for both GC columns.

3.12.4   Complete the header information as described in Section 3.3.  Using
         the same assignment of  first and second GC columns made for Form IV,
         enter the GC column identifier, internal diameter, and date and time
         of analysis(es).  Enter the EPA sample number for the Resolution
         Check Standard.  If simultaneous injections on a single GC are used,
         the EPA sample number may be the same for both Resolution Check
         Standards.   If simultaneous injections are not used, use different
         suffixes to  identify the standards.  Complete the remainder of the
         form using the following instructions.

3.12.4.1       List each analyte, in retention time order, including both
               surrogate compounds.  Thus, the order of analytes in the two
               boxes  on this form will be different due to the dissimilarity
               of the stationary phases of the two GC columns used.  Enter the
               name of each target analyte in the Resolution Check Mixture as
               it appears on Form I PEST.  Spell out the names of the
               surrogates as they appear on Form VII PEST-2.

3.12.4.2       Enter  the 'retention time of each analyte from the analysis in
               the  "RT" column.

3.12.4.3       Calculate the resolution between each pair of analytes.  Enter
               the  resolution between the first and second peaks on  the  line
              • for  the first analyte listed in the box.  Enter the resolution
               between the second and third peaks on the line for the  second
               analyte, and so on, until the resolutions of all possible pairs
                                      B-47                              OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form VII

               of adjacent analytes have been entered.  NOTE:  Only eight of
               the nine resolution fields will be filled.

3.12.4.4       Form VI  (PEST-5, PEST-6 and PEST-7 for each pair of PEM, mid-
               level initial calibration mixture A, and mid-level initial
               calibration mixture B, respectively) shall be used to report
               the percent resolution between each pair of analytes according
           • .   to the definition in Exhibit D (Pesticides).

3.12.4.5       Complete the header information as described in Section 3.3.
               Using the same assignment of first and second GC columns made
               for Form IV, enter the GC column identifier,'internal diameter,
               and date and time of analysis.  Enter the EPA sample number for
               the respective standards.  If simultaneous injections are not
               used, use different suffixes to identify the standards.
               Complete the remainder of the form using the following
               instructions.

3.12.4.5.1        List each analyte, in retention time order, including both
                  surrogate compounds.  Thus, the order of analytes in the two
                  boxes on this form will be different due to the
                  dissimilarity of the stationary phases of the two GC columns
                  used.  Enter the name of each target analyte in the standard
                  as it appears on Form I PEST.  Spell out the names of the
                  surrogates as they appear on Form VII PEST-2.

3.12.4.5.2        Enter the retention time of each analyte from the analysis
                  in the "RT" column.

3.12.4.5.3        Calculate the resolution between each pair of analytes.
                  Enter the resolution between the first and second peaks on
                  the line for the first analyte listed in the box.  Enter the
                  resolution between the second and third peaks on the line
                  for the second analyte, and so on, until the resolutions of
                  all possible pairs of adjacent analytes have been entered.
                  NOTE: The last resolution field will be left blank in each
                  table.

3.13  GC/MS Continuing Calibration Data  (Form VII, VOA-1, VOA-2 and Form VII,
      SV-1, SV-2)

3.13.1   Purpose.  For volatiles and semivolatiles, this form is used to
         report the calibration of the GC/MS system by the analysis of
         specific calibration standards.  Form VII is required for each 12-
         hour time period for both volatile and semivolatile target compound
         analyses.  The Contractor shall analyze calibration standards and
         meet all criteria outlined in Exhibit D for the minimum RRF and
         maximum percent difference between initial and continuing
         calibrations.

3.13.2   -Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Complete the remainder of the form
         using the following instructions.
                                     B-48                              OLM04.2

-------
                                                        Exhibit B -- Section 3
                                                            Forms Instructions
                                                                      Form VII
3.13.2.1    Enter the date and time of the continuing calibration and the
            date(s)  and time(s) of the initial calibration (give inclusive
            dates if the initial calibration is performed over more than one
            date).   Dates shall be entered as MM/DD/YY.  Times shall be
            reported in military time.

3.13.2.2    For vblatiles, enter the purge method, GC column identifier, and
            internal diameter.  For semivolatiles, enter GC column identifier
            and internal diameter.

3.13.2.3    Using the appropriate initial calibration  (volatile or
            semivolatile), enter the average relative response factor  (KRF)
            for each target compound, for each system monitoring compound for
            volatiles, and for each surrogate for semivolatiles.

3.13.2.4    Report the relative response factor  (RRF50) from the continuing
            calibration standard analysis.

3.13.2.5    Calculate the percent difference  (%D) for all compounds.  See
            Exhibit D for equation.  If the %D is greater than 999.9, report
            as 999.9.  If the  %D is less than -99.9, report as -99.9.

3.14  GC/EC Calibration Verification Summary  (Form VII, PEST-1, PEST-2)

3.14.1   Purpose.  Form VII is used to report the results of the Performance
         Evaluation Mixtures  (PEMs) and the mid-point  concentrations of
         Individual Standard Mixtures A and B  that, along with the PEM,
         bracket each 12-hour  period of sample analyses.  The Contractor shall
         submit  Form VII  PEST-1 for each 12-hour  sequence analyzed.  Form VII
         PEST-2  shall be  completed each time the  Individual Standard Mixtures
         are analyzed, for each GC column used.

3.14.2   Instructions.  Complete Form VII PEST-1  and PEST-2 for each standard
         reported on Form VIII PEST.  Complete the header information
         according to the instructions in Section 3.3.  Complete the remainder
         of the  form using the following instructions.

         FORM VII PEST-1

3.14.2.1    Enter the date(s)  of the  initial calibration(s).  Give inclusive
            dates if the  initial calibration  is performed over more than one
            day.  Dates  shall  be entered as MM/DD/YY.

3.14.2.2    Identify the  GC column and internal  diameter in the appropriate
            fields.

3.14.2.3    On Form VII  PEST-1, enter the EPA sample number,  lab sample
            identifier,  and date and  time of analysis  for the  instrument blank
            that preceded the  12-hour sequence  (PIBLK).  For  the PEM that
            initiated or terminated  the 12-hour  sequence  (PEM), enter  the  EPA
            sample number, lab sample identifier, and  date and time of
            analysis.
                                      B-49                               OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form VIII

3.14.2.4    When reporting data for the PEM at the beginning of the initial
            calibration sequence, leave the "EPA Sample No.," "Lab Sample ID,"
            "Date Analyzed," and "Time Analyzed" fields blank for the
            instrument blank  (PIBLK), when no instrument blank is analyzed
            before the PEM.  When reporting all other PEM analyses, the
            instrument blank fields shall be completed.

3.14.2.5    'In the table, report the retention time for each analyte in the
            PEM as well as the retention time windows.

3.14.2.6    For each analyte in the PEM, enter the amount of the analyte found
            in the PEM, in nanograms  (ng) to three decimal places, in the
            "CALC AMOUNT" column.

3.14.2.7    Enter the nominal amount of each analyte in the PEM in the "NOM
            AMOUNT" column.

3.14.2.8    Calculate the percent difference between the calculated amount and
            nominal amount for each analyte according to Exhibit D.  Report
            the values in the "%D" column.  If the %D is greater than 999.9,
            report as 999.9.  If the %D is less than -99.9, report as -99.9.

3.14.2.9    Calculate the percent breakdown for endrin and 4,4'-DDT and the
            combined percent breakdown in the PEM according to Exhibit D.
            Enter the values for the breakdown of endrin and 4,4'-DDT in their
            respective fields immediately under the table.

            FORM VII PEST-2

3.14.2.10   The upper table on Form VII PEST-2 contains the retention time and
            amount data for Individual Standard Mixture A compounds.  The
            lower table contains the data for Mixture B.  Complete the form
            using the instructions in.Sections 3.14.2.1 through 3.14.2.8 for
            Form VII PEST-1.

3.15  Internal Standard Area.and RT Summary  (Form VIII VGA and Form VIII,
      SV-1, SV-2)

3.15.1   Purpose.  This form is used to summarize the peak areas and retention
         times of the internal standards added to all volatile and
         semivolatile samples, including:  dilutions, reanalyses, blanks,
         matrix spikes, and matrix spike duplicates.  The data are used to
         determine when changes in internal standard responses will adversely
         affect quantification of target compounds.  This form shall be
         completed each time a continuing calibration is performed, or when
         samples are analyzed under the same GC/MS instrument performance
         check as an initial calibration.

3.15.2   Instructions.  Complete the header information according to Section
         3.3.  Complete the remainder of the form using the following
         instructions.  If samples 'are analyzed immediately following an
         initial calibration, before another instrument performance check and
         a  continuing calibration, Form VIII shall be completed on the basis
         of the internal standard areas of the 50 ug/L initial calibration


                                     B-50                              OLM04.2

-------
                                                        Exhibit B -- Section 3
                                                            Forms Instructions
                                                                     Form VIII

         standard for volatiles, and the 50 ng initial calibration standard
         for semivolatiles.  Use the date and time of analysis of this
         standard and the lab file identifier and areas in place of those of a
         continuing calibration standard.

3.15.2.1    Enter the date and time of analysis of the continuing calibration
            standard.  The date shall be entered as MM/DD/YY.  The time shall
            be reported as military time.

3.15.2.2    For volatiles, enter the purge method, GC column identifier, and
            internal diameter.  For semivolatiles, enter GC column identifier
            and internal diameter.

3.15.2.3    From the results of the analysis of the continuing calibration
            standard, enter the area measured for each internal standard and
            its retention time  (in decimal minutes) under the appropriate
            column in the "12 HOUR STD" row.

3.15.2.4    For each internal standard listed in Tables 4 and 5, calculate the
            upper limit of the area as the area of the particular standard
            plus 100 percent of its area  (i.e., two times the area in the "12
            HOUR STD" field), and the lower limit of the area as the area of
            the internal standard minus 50 percent of its area  (i.e., one half
            the area in the "12 HOUR STD" field).  Report these values in the
            "UPPER LIMIT" and "LOWER LIMIT" rows, respectively.  Calculate the
            upper limit of the retention time as the retention of the internal
            standard, plus 0.50 minutes  (30 seconds), and the lower limit of
            the retention time as the retention time in the standard minus
            0.50 minutes  (30 seconds).

3.15.2.5    For.each sample,.including dilutions, reanalyses, blanks, matrix
            spikes, and matrix spike duplicates, analyzed under a given
            continuing calibration, enter the EPA sample number and the area
            measured for each internal standard and its retention time.  If
            the internal standard area is outside the upper or lower limits
            calculated in step 4, flag that area with an asterisk  (*).  The
            asterisk shall be placed in the far righthand space of the box for
            each internal standard area, directly under the  "#" symbol.
            Similarly, flag the retention time of any internal standard that
            is outside the limits with an asterisk.

3.15.2.6    Number all pages as described in Section 3.3.

                                    Table 4
                          Volatile Internal Standards


        .Volatile Internal Standards	CAS Number	

        IS1: Bromochloromethane  (BCM)     74-97-5

        IS2: 1,4-Difluorobenzene  (DFB)    540-36-3

        IS3: Chlorbbenzene-d5  (CBZ)     	3114-55-4
                                      B-51                              OLM04.2

-------
Exhibit E -- Section 3
Forms Instructions
Form VIII
                                    Table 5
                        Semivolatiie Internal Standards
          Semivolatile Internal Standards            CAS Number
IS1
IS2
IS3
IS4
IS5
IS6
l,4-Dichlorobenzene-d4 (DCB)
Naphthalene -d8 (NPT)
Acenaphthene-dlO (ANT)
Phenanthrene-dlO (PHN)
Chrysene-dl2 (CRY)
Perylene-dl2 (PRY)
3855-82-1
1146-65-2
15067-26-2
1517-22-2
1719-03-5
1520-96-3
3.16  Pesticide Analytical Sequence  (Form VIII PEST)

3.16.1   Purpose.  This form is used to report the analytical sequence for
         pesticide analysis.  At least one form is required for each GC column
         used for pesticide/Aroclor analyses.

3.16.2   Instructions.  Complete the header  information according to the
         instructions in Section 3.3.  Complete the remainder of the form
         using the following instructions.

3.16.2.1    Enter the date(s) of the initial calibration.  Give inclusive
            dates if the initial calibration is performed over more than one
            day.  Dates shall be entered as MM/DD/YY.

3..16.2.2    Identify the GC column and internal diameter in the appropriate
            fields.

3.16.2.3    At the top of the table, report  the mean retention time for
            tetrachloro-m-xylene  (TCX) and decachlorobiphenyl  (DCB) calculated
            from the initial calibration sequence.

3.16.2.4    For every analysis associated with a particular analytical
            sequence starting with the initial calibration, enter the EPA
            sample number, lab sample identifier, and date and time of
            analysis.  Each sample analyzed  as part of the sequence shall be
            reported on Form VIII PEST even  if it is not associated with the
            SDG.  The Contractor shall use ZZZZZ as the EPA sample number to
            distinguish all samples that are not part of the SDG being
            reported.

3.16.2.5    Report the retention time of .the surrogates for each analysis in
            the "TCX RT" and "DCB RT" columns.  All sample analyses shall be
            bracketed by acceptable analyses of instrument blanks, a PEM, and
            Individual Standard Mixtures A and B.  Given the fact that the
            initial calibration may remain valid for some time  (see Exhibit
            D), it is only necessary to report the data from 12-hour periods
            when samples, dilutions, reanalyses, matrix spike, matrix spike
            duplicate, blanks, or multicomponent analytes for the 72 hour


                                     B-52                              OLM04.2

-------
                                                          Exhibit B--Section 3
                                                            Forms Instructions
                                                                       Form IX

            confirmation requirement in an SDG were analyzed.   All data
            necessary to demonstrate compliance with the requirements
            specified in Exhibit D-Pest Section 9.3 must be reported.  The
            Contractor shall submit Form VIII for the initial  calibration
            sequence and forms that include the PEMs and Individual Standard
            Mixtures that bracket any and all samples in the SDG.  While the
            data for time periods between the initial calibration and samples
            in the SDG are not a routine deliverable, the data shall be
            available as requested  (e.g., at on-site evaluations).  Non-EPA
            samples or samples from SDGs not being reported shall be numbered
            ZZZZZ.

3.16.2.6    Flag all those values which do not meet the contract requirements
            by entering an asterisk (*) in the "RT" column, under the "#"
            symbol.  If the retention time cannot be calculated due to
            interfering peaks, leave the "RT" column blank for that surrogate,
            enter an asterisk in the last column, and document the problem in
            the SDG Narrative.
3.16.2.7    If more than a single copy of Form VIII PEST is required, enter
            the same header information on all subsequent pages for that GC
            column and instrument, and number each page as described in
            Section 3.3.

3.17  Pesticide Cleanup Summary  (Form IX, PEST-1, PEST-2)

3.17.1   Purpose.  This form summarizes the results of the checks performed
         for both cleanup procedures employed during the preparation of
         pesticide extracts for analysis.  Form IX PEST-1 is used to report
         the results of the check of the Flori'sil cartridges used to process
         all sample extracts and to associate the lot of cartridges with
         particular sample results so that problems with a particular
         cartridge lot may be tracked across all associated samples.  Form IX
         PEST-2 summarizes the results of the calibration verification of the
         Gel Permeation Chromatography  (GPC) device that shall be used to
         process all soil sample extracts for pesticide/Aroclor analyses.

3.17.2   Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Enter the Case number and SDG number
         for the current data package, regardless of the original Case for
         which the cartridge check was performed.  Complete the remainder of
         the form using the following instructions.

            FORM IX PEST-1

3.17.2.1    Enter the Florisil cartridge lot number.

3.17.2.2    Enter the date the Florisil  cartridge check solution was analyzed
            in the  "Date of Analysis" field.  The date shall be entered as
            MM/DD/YY.

3.17.2.3    Complete the  "GC Column" and "ID" fields for the two GC  columns
            used to analyze the  samples, including blanks, matrix spikes, and


                                     B-53                              OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form X

            matrix spike duplicates.  Report all results from either GC column
            1 or GC column 2.

3.17.2.4    In the first table, enter the amount of spike added and spike
            recovered in nanograms  (ng) for each analyte.

3.17.2.5    Calculate the percent recovery to the nearest whole percent, and
            enter the number in the "% REC" field.  Flag each spike recovery
            outside the QC limits (shown on the form) with an asterisk  (*) .
            The asterisk shall be placed in the last space in the "% REC"
            column, under the  "#" symbol.

3.17.2.6    In the second table, complete the "EPA Sample No.," the "Lab
            Sample ID," and  "Date Analyzed" fields for each sample and blank
            that were cleaned up using this lot of Florisil cartridges.

3.17.2.7    Number the pages as described in Section 3.3.

            FORM IX PEST-2

3.17.2.8    On Form IX PEST-2, enter an identifier for the GPC column and the
            analysis date of calibration verification in the appropriate
            fields.

3.17.2.9    Complete the "GC Column" and "ID" fields as on Form IX PEST-1 for
            florisil.  Report all results from a single column.

3.17.2.10   For each of the pesticide matrix spike compounds listed in the
            first table, enter the amount of the spike added to the GPC column
            and the amount recovered, in nanograms (ng).

3.17.2.11   Calculate the percent recovery of each analyte, and enter these
            values on the form, to the nearest percent.  Compare the
            recoveries to the QC limits shown on the form, and flag all those
            values outside the limits with an asterisk (*) in the "% REC"
            column under the "#" symbol.

3.17.2.12   For each sample  in the data package that was subjected to GPC
            under this calibration verification, enter the EPA sample number,
            lab sample identifier, and the date of both analyses in the second
            table.

3.17.2.13   If more than one copy of Form IX PEST-2 is required, number all
            pages as described in Section 3.3.

3.18  Pesticide/Aroclor Identification  (Form X, PEST-1, PEST-2)

3.18.1   Purpose.  This form summarizes the quantitations of all target
         pesticides/Aroclors detected in a given sample.  It reports the
         retention times of  the compound on both columns on which it was
         analyzed, as well as the retention time windows of the standard for
         that compound on both of these columns.  In addition, it is used to
         report the concentration determined from each-GC column, and the
         percent difference between the two quantitative results.  Separate


                                     B-54                              OLM04.2

-------
                                                          Exhibit B--Section 3
                                                            Forms Instructions
                                                                        Form X

         forms are used for single component analytes and multicomponent
         analytes.

         Form X is required for each sample, including dilutions and
         reanalyses, blank, matrix spike,  and matrix spike duplicate in which
         compounds listed in Exhibit C (Pesticides/Aroclors) are reported on
         Form I.  Do not generate a Form X for pesticide instrument blanks.

3.18.2   Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Complete the remainder of the form
         using the following instructions.

3.18.2.1       Enter the date(s) of analysis.  Dates shall be entered as
               MM/DD/YY.

3.18.2.2       Enter the GC column and internal diameter for each of the two
               columns.

3.18.2.3       For each single component pesticide positively identified,
               enter the name of the compound in the "ANALYTE" column as it
               appears on Form I.

3.18.2.4       Enter the retention times on each column of the compounds
               detected in the sample next to the appropriate column
               designation  (1 or 2).

3.18.2.5       Enter the retention time windows on each column from the
               initial calibration standard. . These data shall correspond with
               those on Form VI and shall be entered in a similar manner.  The
               lower value is entered under the "FROM" column, the upper value
               under the "TO" column.

3.18.2.6       Enter the concentration calculated from each GC column under
               the  "CONCENTRATION" column.  Although the units are the same as
               those used on Form I, ug/L for water samples and ug/Kg for soil
               samples, do not enter any units on Form X.

3.18.2.7       Calculate the percent difference between the concentrations
               entered on  this form.  See Exhibit D for equation, and report
               to a tenth of a percent in the  "%D" column.  If the  %D is
               greater than 999.9, report it as 999.9.

3.18.2.8       The  lower of the  two concentrations is reported on Form I for
               each pesticide compound.  The lower concentration is used
               because, if present,  coeluting  interferences are  likely to
                increase the calculated concentration of any target  compound.
                If the percent difference between the calculated  concentrations
                is greater  than 25.0 percent, flag the concentration on Form I,
               as described previously.  This  will alert the  data user to  the
               potential problems  in quantitating this analyte.

3.18.2.9        If more pesticide compounds  are identified in  an  individual
                sample  than can be  reported  on  one Form X, complete  as many
                additional  copies of  Form X  as  necessary, duplicating all


                                     B-55                              OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form DC-1

               header information and numbering the pages as described in
               Section 3.3.

3.18.2.10      Report multicomponent analytes detected in samples on Form X
               PEST-2.   Complete the header information and GC column fields
               as described above.  For multicomponent analytes, it is
               necessary to report the retention time and concentration of
               each peak chosen for quantitation in the target analyte in a
               fashion similar to that for single component pesticides.  The
               concentrations of all peaks quantitated (three are required, up
               to five may be used) are averaged to determine the mean
               concentration.  Report the lower of the two mean concentrations
               on Form I.  Flag this value if the mean concentrations from the
               two GC columns differ by more than 25 percent, as described
               previously.

3.18.2.11      If more multicomponent compounds,are identified in an
               individual sample than can be reported on one Form X, complete
               as many additional copies of Form X. as necessary, duplicating
               all header information and numbering the pages as described in
               Section 3.3.

3.19  Sample Log-In Sheet  (Form DC-1)

3.19.1   Purpose.  This form is used to document the receipt and inspection of
         sample containers and samples.  One original of Form DC-1 is required
         for each sample shipping container  (only the hardcopy form is-
         required).  If the samples in a single sample shipping container are
         assigned to more than one SDG, the original Form DC-1 shall be placed
         with the deliverables for the SDG of the lowest alpha-numeric number,
         and a copy of Form DC-1 shall be placed with the deliverables for the
         other SDGs.  The copies shall be identified as "copy(ies)", and the
         location of the original shall be noted on the copies.

3.19.2   Instructions

3.19.2.1    Sign and date the airbill  (if present).

3.19.2.2    Complete the header information on the form,  including the log-in
            date.

3.19.2.3    Examine the shipping container and record the presence/absence of
            custody seals and their condition  (e.g., intact, broken) in item
            1.

3.19.2.4    Record the custody seal numbers in item 2.

3.19.2.5    Open the container, remove the enclosed sample documentation, and
            record the presence/absence of chain-of-custody record(s), SMO
•            forms (e.g., Traffic Reports, Packing Lists), and airbills or
            airbill stickers in items 3-5.  Specify if there is an airbill
            present or an airbill sticker in item 5.  Record the airbill or
            sticker number in item 6.
                                     B-56                              OLM04.2

-------
                                                            Exhibit B--Section 3
                                                              Forms Instructions
                                                                       Form DC-2

  3.19.2.6    Remove the samples from the shipping container(s),  examine the
              samples and the sample tags (if present),  record the condition of
              the sample bottles (e.g., intact, broken,  leaking), and presence
|  .            or absence of sample tags in items 7 and 8.

|  '3.19.2.7    Record the cooler temperature in item 9.

\  3.19.2.8    Review the sample shipping documents and compare the information
              recorded on all the documents and samples and circle the
|              appropriate answer in item 10.

|  3.15.2.9    Record the date and time of cooler receipt at the laboratory in
|              items 11 and 12.

|  3.19.2.10   If there are no problems observed during receipt, sign and date
              (include the time) Form DC-1, the chain-of-custody record, the
              Traffic Report, and write the sample numbers on  Form DC-1 in the
              "EPA Sample #" column.

|  3.19.2.11   Record the appropriate sample tags and  assigned  laboratory
              numbers, if applicable.

|  3.19.2.12   Any comments should be made in the "Remarks" column.

|  3.19.2.13   Record the fraction designation  (if appropriate) and the specific
              area designation  (e.g.,  refrigerator number) in  the "Sample
              Transfer" block.  Sign and date  the "Sample Transfer" block.

|  3.19.2.14   Cross out unused columns and  spaces.

|  3.19.2.15   If there are problems observed during receipt or an answer marked
              with an asterisk  (e.g.,  "absent*") was  circled,  contact SMO and
              document the contact as  well  as  resolution of the problem on a CLP
              Communication Log.  Following resolution, sign and date the forms
              and note, where appropriate,  the resolution of the problem.

  3.20  Document Inventory Sheet  (Form DC-2)

  3.20.1   Purpose.  The Document Inventory Sheet  (Form DC-2)  is used to record
           both the inventory of Complete SDG  File  (CSF) documents and the
           number of documents in the  original sample data package which is sent
           to the EPA Region.

  3.20.2   Instructions

  3.20.2.1    Organize all EPA CSF documents as described in Exhibit B, Sections
              II and III.  Assemble the documents in  the order specified on Form
              DC-2 and Sections  II and III, and stamp each page with a
              consecutive number; however,  do  not number Form  DC-2.   Inventory
              the CSF by  reviewing the document numbers and recording page
              number ranges  in the columns  provided on Form DC-2.  The
              Contractor  shall verify  and record in the "Comments" section on
              Form DC-2 all  intentional gaps in the page numbering sequence  (for
              example,  "page numbers not used, XXXX - XXXX, XXXX  - XXXX."  If


                                        B-57                             OLM04.2

-------
Exhibit B--Section 3
Forms Instructions
Form DC-2

            there are no documents for a specific document type, enter a "NA"
            in the empty space.

3.20.2.2    Certain laboratory-specific documents related to the CSF may not
            fit into a clearly defined category.  The Contractor shall review
            Form DC-2 to determine if it is most appropriate to place them
            under categories 7, 8, 9, or 10.  Category 10 should be used if
            there is no appropriate previous category.  These types of
            documents should be described or listed in the blanks under each
            appropriate category on Form DC-2.

3.20.2.3    If it is necessary to insert new or inadvertently omitted
            documents, the Contractor shall identify the documents with unique
            accountable numbers and record the unique accountable numbers and
            the locations of the documents in the CSF (in the "Other Records",
            section on Form DC-2).
                                     B-58                              OLM04.2

-------
                                                        Exhibit B --  Section 4
                                                          Data Reporting Forms
4.0   DATA REPORTING FORMS

      The data reporting forms are shown on the following pages.
                                      B-59                              OLM04.2

-------
                                       1A
                     VOLATILE ORGANICS ANALYSIS DATA SHEET
                                           EPA SAMPLE NO.
Lab Name:

Lab Code:
               Contract:
Matrix:  (soil/water).

Sample wt/vol: 	
Case No.:
   .(g/mL).
  SAS No.:
              SDG No.:
Level:  (low/med).
% Moisture: not dec..

GC Column: 	
Soil Extract Volume:
      ID:
(mm)
         Lab Sample ID:

         Lab File ID:
                           Date Received:

                           Date Analyzed:
Dilution Factor:
                           Soil Aliquot Volume:
                                   .(uL)
   CAS NO.   COMPOUND
                              CONCENTRATION UNITS:
                              (ug/L or ug/Kg)	
75-71-8
74-87-3
75-01-4
74-83-9
75-00-3
75-69-4
75-35-4
76-13-1
67-64-1
75-15-0
79-20-9
75-09-2
156-60-5
1634-04-4
75-34-3
156-59-2
78-93-3
67-66-3
71-55-6
110-82-7
56-23-5
71-43-2
107-06-2
Di chlorodi f luorome thane
Chloromethane
Vinyl Chloride
Bromomethane
Chloroethane
Trichlorof luoromethane
1, 1-Dichloroethene
1,1, 2-Trichloro-l, 2, 2-trif luoroethane
Acetone
Carbon Disulfide
Methyl Acetate
Methylene Chloride
trans-1, 2-Dichloroethene
Methyl tert-Butyl Ether
1 , 1-Dichloroethane
cis-1, 2-Dichloroethene
2-Butanone
Chloroform
1,1, 1-Trichloroethane
Cyclohexane
Carbon Tetrachloride
Benzene
1 , 2 -Dichloroethane














































                                  FORM I  VOA-1
                                                                        OLM04.2

-------
Lab Name:

Lab Code:
                   IB
 VOLATILE ORGANICS ANALYSIS DATA SHEET

	   Contract:  	

 Case No. :  	 SAS No.  : 	
                                                                EPA SAMPLE NO.
                         SDG No.:
Matrix:  (soil/water)	

Sample wt/vol: 	(g/mL).
                             Lab Sample ID:

                             Lab File ID:
Level:  (low/med).
% Moisture: not dec..

GC Column: 	
                             Date Received:

                             Date Analyzed:
     ID:
(mm)
                             Dilution Factor:
Soil Extract Volume:
    CAS NO.   COMPOUND
                             Soil Aliquot Volume: 	

                               CONCENTRATION UNITS:
                                (ug/L  or  ug/Kg)	
                                      .(UL)
79-01-6
108-87-2
78-87-5
75-27-4
10061-01-5
108-10-1
108-88-3
10061-02-6
79-00-5
127-18-4
591-78-6
124-48-1
106-93-4
108-90-7
100-41-4
1330-20-7
100-42-5
75-25-2
98-82-8
79-34-5
541-73-1
106-46-7
95-50-1
96-12-8
120-82-1
Trichloroethene
Me thy 1 eye lohexane
1 , 2 -Dichlorbpropane
Bromodichloromethane
cis-1, 3-Dichloropropene
4 -Methyl - 2 -pentanone
Toluene
trans-1, 3-Dichloropropene
1,1, 2-Trichloroethane
Tetrachloroethene
2-Hexanone
Dibromochloromethane
1 , 2 - D ibromoe thane
Chlorobenzene
Ethylbenzene
Xylene (total)
Styrene
Bromoform
Isopropylbenzene
1 , 1 , 2 , 2 - Tetrachloroethane
1 , 3 -Dichlorobenzene
1,4-Dichlorobenzene
1 , 2 -Dichlorobenzene
1, 2-Dibromo-3-chloropropane
1,2, 4 -Trichlorobenzene


















































                                  FORM I VOA-2
                                                     OLM04.2

-------
Lab Name:
Lab Code:
                                      1C
                   SEMIVOLATILE ORGANICS ANALYSIS DATA SHEET
                  Contract:
                                            EPA SAMPLE NO.
Case No.:
SAS No.:
                                                              SDG No.
Matrix: (soil/water).

Sample wt/vol: 	
      .(g/mL).
         Lab  Sample  ID:

         Lab  File  ID:
Level: (low/med)

% Moisture: 	
  Decanted:(Y/N)
Concentrated Extract Volume:	(uL)

Injection Volume:	(uL)
         Date Received:

         Date Extracted:

         Date Analyzed:
                            Dilution Factor:
GPC Cleanup:  (Y/N).
         PH:.
       CAS NO.  COMPOUND
          Extraction:  (Type).
        CONCENTRATION  UNITS:
        (ug/L  or  ug/Kg)	
100-52-7
108-95-2
111-44-4
95-57-8
95-48-7
108-60-1
98-86-2
106-44-5
621-64-7
67-72-1
98-95-3
78-59-1
88-75-5
105-67-9
111-91-1
120-83-2
91-20-3
106-47-8
87-68-3
105-60-2
59-50-7
91-57-6
77-47-4
88-06-2
95-95-4
92-52-4
91-58-7
88-74-4
131-11-3
606-20-2
208-96-8
99-09-2
83-32-9
Benzaldehyde
Phenol .
bis (2-Chloroethyl) ether
2-Chlorophenol
2 -Methylphenol
2,2' -oxybis (1-Chloropropane)
Acetophenone
4 -Methylphenol
N-Nitroso-di-n-propylamine
Hexachloroethane • . .
Nitrobenzene
Isophorone
2-Nitrophenol
2 , 4 -Dimethylphenol
bis (2 -Chloroethoxy) methane
2 , 4 -Dichlorophenol
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
Caprolactam
4 -Chloro- 3 -methylphenol
2 -Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6 -Trichlorophenol
2,4 , 5-Trichlorophenol
1,1' -Biphenyl
2-Chloronaphthalene
2-Nitroaniline
Dimethylphthalate
2, 6-Dinitrotoluene
Acenaphthylene
3-Nitroaniline
Acenaphthene


























,





.

































                                  FORM I SV-1
                                                                        OLM04.2

-------
                                        ID
                     SEMIVOLATILE  ORGANICS ANALYSIS  DATA SHEET
                                                                  EPA SAMPLE NO.
  Lab Name:
  Lab Code:
                 Contract:
                      Case No.:
                 SAS No.:
               SDG NO.:
  Matrix: (soil/water).

  Sample 'wt/vol:  	
  .(g/mL).
  Leve1:  (1ow/me d)
  % Moisture:
Decanted: (Y/N).
  Concentrated Extract Volume:
  Injection Volume:
|  GPC Cleanup:  (Y/N).
    pH:
                 .(uL)
  Lab Sample ID:	

  Lab File ID:  	

  Date Received:

  Date Extracted:

  Date Analyzed:
                          Dilution  Factor:
         CAS NO.  COMPOUND
  Extraction: (Type).
CONCENTRATION UNITS:
(ug/L or ug/Kg)	
51-28-5
100-02-7
132-64-9
121-14-2
84-66-2
86-73-7
7005-72-3
100-01-6
534-52-1
86-30-6
101-55-3
118-74-1
1912-24-9
87-86-5
85-01-8
120-12-7
86-74-8
84-74-2
206-44-0
129-00-0
85-68-7
91-94-1
56-55-3
218-01-9
117-81-7
117-84-0
205-99-2
207-08-9
50-32-8
193-39-5
53-70-3
191-24-2
2 , 4 -Dinitrophenol
4 -Nitrophenol
Dibenzofuran
2 , 4 -Dinitrotoluene
Diethylphthalate
Fluorene
4 -Chlorophenyl -phenylether
4-Nitroaniline
4, 6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine (1)
4 -Bromophenyl-phenylether
Hexachlorobenzene
Atrazine
Pentachlorophenol
Phenanthrene
Anthracene
Carbazole
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Benzo (a) anthracene
Chrysene
bis ( 2 - Ethylhexyl ) phthalate
Di -n-octylphthalate
Benzo (b) f luoranthene
Benzo (k) f luoranthene
Benzo (a) pyrene
.Indeno (1,2, 3 -cd) pyrene
Dibenzo (a, h) anthracene
Benzo (q,h, i)perylene



















'












































   (1)  Cannot be separated from Diphenylamine
                                    FORM  I SV-2
                                                  OLM04.2

-------
Lab Name:
Lab Code:
                                      IE
                    PESTICIDE ORGANICS ANALYSIS DATA SHEET
               Contract:
                                            EPA SAMPLE NO.
Case No. :
SAS No.:
SDG No.
Matrix:  (soil/water)	

Sample wt/vol: 	(g/mL)	

% Moisture: 	 Decanted:  (Y/N).

Extraction:  (Type) 	
                          Lab Sample ID:

                          Lab File ID:
Concentrated Extract Volume: _

Injection Volume: 	(uL)

GPC Cleanup:  (Y/N)	  pH:
                  .(uL)
   Date Received:

   Date Extracted:

   Date Analyzed:
                          Dilution Factor:
       CAS NO.  COMPOUND
                          Sulfur Cleanup:  (Y/N)
                          CONCENTRATION UNITS:
                          (ug/L or ug/Kg)
                                Q
319-84-6
319-85-7
319-86-8
58-89-9
76-44-8
309-00-2
1024-57-3
959-98-8
60-57-1
72-55-9
72-20-8
33213-65-9
72-54-8
1031-07-8
50-29-3
72-43-5
53494-70-5
7421-93-4
5103-71-9
5103-74-2
8001-35-2
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
alpha -BHC
beta-BHC
delta-BHC
gamma -BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4, 4 '-DDE
Endrin
Endosulfan II
4/4 '-ODD
Endosulfan sulfate
4, 4 '-DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha-Chlordane
gamma - Chlordane
Toxaphene
Aroclor-1016
Aroclor-1221
Aroclor-1232
Aroclor-1242
Aroclor-1248
Aroclor-1254
Aroclor-1260
























,































                                  FORM I -PEST
                                                    OLM04.2

-------
                                      IF
                     VOLATILE ORGANICS ANALYSIS DATA SHEET
                       TENTATIVELY IDENTIFIED  COMPOUNDS
                                              EPA SAMPLE NO.
Lab Name:

Lab Code:
Case No.:
Contract:

SAS No.:
SDG No.:
Matrix:  (soil/water).

Sample wt/vol: 	
   .(g/mL).
     Lab Sample ID:

     Lab File ID:
Level:  (low/med).
% Moisture: not dec.	

GC Column: 	  ID:

Soil Extract Volume:  	

Number TICs found: 	
                         Date Received:

                         Date Analyzed:
            .(mm)
     Dilution Factor:
                         Soil Aliquot Volume:

                         CONCENTRATION UNITS:
                          (ug/L or ug/Kg)
                                    .(uL)
CAS NUMBER
1.
2.
3.
4 .
5.
6.
7.
8.
9.
10.
11.
12.
13.
14 .
15.
16.
17.
18.
19.
20.
21.
22.
23.
24 .
25.
26.
27.
28.
29.
30'.
COMPOUND NAME






























RT






























EST . CONC .






























Q






























                                 FORM I VOA-TIC
                                                     OLM04.2

-------
  Lab Name:

  Lab Code:
                    1G
 SEMIVOLATILE ORGANICS ANALYSIS DATA SHEET
     TENTATIVELY IDENTIFIED COMPOUNDS
	  Contract:     .	
                                                                   EPA SAMPLE NO.
 Case No.:
                     SAS  NO.:
            SDG No.:
  Matrix: (soil/water).

  Sample wt/vol:  	
     .(g/mL).
Lab Sample ID:

Lab File ID:
  Level: (low/med).

  % Moisture: 	
     Decanted:  (Y/N).
  Concentrated Extract Volume: 	

  Injection Volume: 	(pL)

|  GPC Cleanup: (Y/N)	      pH:

 • Number TICS found: 	
                  .(uL)
Date Received:

Date Extracted:

Date Analyzed:
                             Dilution Factor:
                             Extraction: (Type).
                             CONCENTRATION UNITS:
                             (ug/L or ug/Kg)
CAS NUMBER
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
. 14'.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
COMPOUND NAME


-



























RT






























EST. CONC.












-•

















Q


'















'











                                   FORM I SV-TIC
                                                      OLM04.2

-------
                                      2A
              WATER VOLATILE SYSTEM MONITORING COMPOUND RECOVERY
Lab Name:

Lab Code:
Case No.
                    Contract:
SAS No.:
SDG No.

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
2.1
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.











'


.
^














SMC1
(TOL) #






























SMC2
(BFB) #






























SMC3
(DCE) #






























OTHER






























TOT
OUT






























           SMC1  (TOL) = Toluene-d8
           SMC2  (BFB) = Bromofluorobenzene
           SMC3  (DCE) = l,2-Dichloroethane-d4

           # Column to be used to flag recovery values

           * Values outside of contract required QC  limits
                                           QC LIMITS
                                            (88-110)
                                            (86-115)
                                            (76-114)
 page
             of
                                  FORM  II VOA-1
                                                    OLM04.2

-------
                                      2B
               SOIL VOLATILE SYSTEM MONITORING COMPOUND RECOVERY
Lab Name:

Lab Code:
Case No.:
Contract:

SAS No.:
                                                        SDG No.:
Level:  (low/med).
01
02
03
04
05
06
07
08 .
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























SMC1
(TOL) #





















_, .








SMC2
(BFB) #






























SMC3
(DCE) #






























OTHER






























TOT
OUT






























QC LIMITS
SMC1 (TOL) = Toluene-d8 (84-138)
SMC2 (BFB) = Bromofluorobenzene (59-113)
SMC3 (DCE) = l,2-Dichloroethane-d4 (70-121)
           # Column to be used to flag recovery values

           * Values outside of contract required QC limits
page
            .of
                                 FORM II VOA-2
                                                    OLM04.2

-------
Lab Name:

Lab Code:
                 - 2C
WATER SEMIVOLATILE  SURROGATE  RECOVERY

	   Contract: 	

            	   SAS No.: 	
                    Case  No.:
SDG No.:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























SI
(NBZ)#






























S2
(FBP)#






























S3
(TPH)#






























S4
(PHL)#






























S5
(2FP)#






























S6
(TBP)#






























S7
(2CP)#






























S8
(DCB)#





-
























TOT
OUT






























 page
QC LIMITS
SI (NBZ)
S2 (FBP)
S3 (TPH)
S4 (PHL)
S5 (2FP)
S6 (TBP)
S7 (2CP)
S8 (DCB)
# Column
* Values
= Nitrobenzene-d5
= 2 - Fluorobiphenyl
= Terphenyl-dl4
= Phenol -d5
= 2-Fluorophenol
= 2,4,6- Tr ibromophenol
= 2-Chlorophenol-d4
= 1, 2-Dichlorobenzene-d4
(35-114)
(43-116)
(33-141)
(10-110)
(21-110)
(10-123)
(33-110)
(16-110)






(advisory)
(advisory)
to be used to flag recovery values
outside of contract required QC
limits

D Surrogate diluted out
of



                                  FORM II SV-1
                                                                        OLM04.2

-------
Lab Name:
Lab Code:
                   2D
 SOIL SEMIVOLATILS SURROGATE RECOVERY

 	   Contract:  	
Case No.:
SAS No.
                 SDG No.:
Level: (low/med).
EPA
SAMPLE NO.






























SI
(NBZ)#






























S2
(FBP)#






























S3
(TPH) #






























S4
(PHL)#






























S5
(2FP)#






























S6
(TBP)#






























S7
(2CP)#






























S8
(DCB)#






























TOT
OUT






























01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
           SI (NBZ)  = Nitrobenzene-d5
           S2 (FBP)  = 2-Fluorobiphenyl
           S3 (TPH)  = Terphenyl-dl4
           S4 (PHL)  = Phenol-d5
           S5 (2FP)  = 2-Fluorophenol
           S6 (TBP)  = 2,4,6-Tribromophenol
           S7 (2CP)  = 2-Chlorophenol-d4
           S8 (DCB)  = l,2-Dichlorobenzene-d4
                                   QC LIMITS
                                   (23-120)
                                   (30-115)
                                   (18-137)
                                  • (24-113)
                                   (25-121)
                                   (19-122)
                                   (20-130)
                                   (20-130)
                         (advisory)
                         (advisory)
           #  Column to be used to flag recovery values
           *  Values outside of contract required QC limits
           D  Surrogate diluted out
page
            of
                                  FORM  II  SV-2
                                                                        OLM04.2

-------
                                      2E
                      WATER PESTICIDE SURROGATE RECOVERY
Lab Name:
                    Contract:
Lab Code:
GC Column(1):
 Case No.:

	 ID:
     SAS No.:
SDG NO.:
(mm)  GC Column (2) :
       ID:
.(mm)

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























TCX 1
%REC #






























TCX 2
%REC #






























DCB 1
%REC #






























DCB 2
%REC tt






























OTHER
(1)






























OTHER
(2)






























TOT
OUT






























           TCX   =  Tetrachloro-m-xylene
           DCB   =  Decachlorobiphenyl
                                    QC LIMITS
                                    (30-150)
                                    (30-150)
           #  Column to be used  to  flag  recovery values
           *  Values outside  of  QC  limits
           D  Surrogate diluted  out
 page
             of
                                 FORM II.PEST-1
                                                                        OLM04.2

-------
                                      2F
                       SOIL  PESTICIDE  SURROGATE  RECOVERY
Lab Name :
Lab Code:
GC Column (1) :

Case No. :
ID:
Contract :
SAS No . : SDG No . :
(mm) GC Column (2) : ID:


(mm)
01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























TCX 1
%REC #






























TCX 2
%REC #






























DCB 1"
%REC #






























DCB 2
%REC #






























OTHER
(1)






























OTHER
(2)






























TOT
OUT






























           TCX   =   Tetrachloro-m-xylene
           DCB   =   Decachlorobiphenyl
QC LIMITS
(30-150)
(30-150)
           #   Column to be used to flag recovery values
           *   Values outside of QC limits
           D   Surrogate diluted out
page
            of
                                FORM II PEST-2
                 OLM04.2

-------
                                      3A
         WATER VOLATILE MATRIX  SPIKE/MATRIX  SPIKE DUPLICATE RECOVERY
Lab Name:

Lab Code:
                   Case No.:
Matrix Spike - EPA Sample No.:
  Contract:

SAS No.:
SDG No.:
COMPOUND
1, 1-Dichloroethene
Trichloroethene
Benzene
Toluene
Chlorobenzene
SPIKE
ADDED
(ug/L)





SAMPLE
CONCENTRATION
' (ug/L)





MS
CONCENTRATION
(ug/L)





MS
%
REC #





QC
LIMITS
REC.
61-145
71-120
76-127
76-125
75-130
COMPOUND
1, 1-Dichloroethene
Trichloroethene
Benzene
Toluene
Chlorobenzene
SPIKE
ADDED
(ug/L)





MSD
CONCENTRATION
(ug/L)





MSD
%
REC #





%
RPD #





QC
RPD
14
14
11
13
13
LIMITS
REC.
61-145
71-120
76-127
76-125
75-130
 #  Column to  be  used to flag recovery and RPD values with an asterisk

 *  Values outside  of QC limits

 RPD: 	  out of 	 outside limits
 Spike  Recovery: 	 out of 	 outside  limits
 COMMENTS:
                                 FORM III VOA-1
                                   OLM04.2

-------
                                      3B
          SOIL VOLATILE MATRIX SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
Lab Name:

Lab Code:
Case No.:
Matrix Spike - EPA Sample No.:
                    Contract:

                  SAS No.:
                                                       SDG No.:
                             Level:(low/med).
COMPOUND
1, 1-Dichloroethene
Trichloroethene
Benzene
Toluene
Chlorobenzene
SPIKE
ADDED
(ug/Kg)





SAMPLE
CONCENTRATION
(ug/Kg)





MS
CONCENTRATION
(ug/Kg)





MS
%
REC #





QC
LIMITS
REC.
59-172
62-137
66-142
59-139
60-133
COMPOUND
1, 1-Dichloroethene
Trichloroethene
Benzene
Toluene
Chlorobenzene
SPIKE
ADDED
(ug/Kg)





MSD
CONCENTRATION
(ug/Kg)





MSD
%
REC #





%
RPD #





QC
RPD
22
24
21
21
21
LIMITS
REC.
59-172 .
62-137
66-142
59-139
60-133
# Column to be used to flag recovery and RPD values with an asterisk

* Values outside of 'QC limits
RPD: 	 out of
Spike Recovery: 	
COMMENTS:
   	 outside limits
   out of 	 outside limits
                                FORM  III VOA-2
                                                                        OLM04.2

-------
                                      3C
       WATER SEMIVOLATILE MATRIX SPIKE/MATRIX  SPIKE  DUPLICATE RECOVERY
Lab Name:

Lab Code:
                    Case No.:
                                      Contract:

                                      SAS No.:
                                 SDG No.:
Matrix Spike - EPA Sample No.:
COMPOUND
Phenol
2 - Chlorophenol
N-Nitroso-di-n-prop. (1)
4 -Chloro- 3 -methylphenol
Acenaphthene
4 -Nitrophenol
2 , 4 -Dinitrotoluene
Pentachlorophenol
Pyrene
SPIKE
ADDED
(ug/L)









SAMPLE
CONCENTRATION
(ug/L)









MS
CONCENTRATION
(ug/L)









MS
%
REC #









QC
LIMITS
REC.
12-110
27-123
41-116
23-97
46-118
10-80
24-96
9-103
26-127
COMPOUND
Phenol
2 -Chlorophenol
N-Nitroso-di-n-prop. (1)
4 - Chloro - 3 -methylphenol
Acenaphthene
4 -Nitrophenol
2 , 4 -Dinitrotoluene
Pentachlorophenol
Pyrene
SPIKE
ADDED
(ug/L)









MSD
CONCENTRATION
(ug/L)









MSD
%
REC #









%
RPD #









QC L]
RPD
42
40
38
42
31
' 50
38
50
31
EMITS
REC.
12-110
27-123
41-116
23-97
46-118
10-80
24-96
9-103
26-127
 (1) N-Nitroso-di-n-propylamine


 #  Column  to be  used to flag recovery and  RPD  values  with an asterisk

 *  Values  outside of QC limits
             out of
RPD:	
Spike Recovery:

COMMENTS:  	
	 outside limits
 out of  	  outside  limits
                                 FORM .III SV-1
                                                                       OLM04.2

-------
                                      3D
        SOIL SEMIVOLATILE MATRIX SP.IKE/MATRIX SPIKE DUPLICATE RECOVERY
Lab Name:

Lab Code:
Case No.:
Matrix Spike - EPA Sample No.:
                   Contract:

                   SAS No.:
SDG No.:
                   Level:(low/med)
COMPOUND
Phenol
2-Chlorophenol
N-Nitroso-di-n-prop. (1)
4-Chloro-3-methylphenol
Acenaphthene
4 -Nitrophenol
2 , 4 -Dinitrotoluene
Pentachlorophenol
Pyrene
SPIKE
ADDED
(ug/Kg)









SAMPLE
CONCENTRATION
(ug/Kg)









MS
CONCENTRATION
(ug/Kg)




-




MS
%
REC #









QC
LIMITS
REC.
26-90
25-102
41-126
26-103
31-137
11-114
28-89
17-109
35-142
COMPOUND
Phenol
2 -Chlorophenol
N-Nitroso-di-n-prop. (1)
4-Chloro-3-methylphenol
Acenaphthene
4 -Nitrophenol
2 , 4 -Dinitrotoluene
Pentachlorophenol
Pyrene
SPIKE
ADDED
(ug/Kg)









MSD
CONCENTRATION
(ug/Kg)









MSD
%
REC #









%
RPD #









QC L
PRO
35
50
38
33
19
50
47
47
36
IMITS
REC. '
26-90
25-102
41-126
26-103
31-137
11-114
28-89
17-109
35-142
 (1) N-Nitroso-di-n-propylamine


# Column to be used to flag recovery and RPD values with an asterisk

* Values outside of QC limits

RPD: 	 out of 	 outside limits
Spike Recovery: 	out of 	 outside limits
COMMENTS:
                                 FORM III SV-2
                                                    OLM04.2

-------
                                      3E
         WATER PESTICIDE MATRIX SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
Lab Name:

Lab Code:
                   Case No. -.
Matrix Spike - EPA Sample No.:
  Contract:

SAS No.:
SDG No.
COMPOUND
gamma -BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4, 4 '-DDT
SPIKE
ADDED
(ug/L)






SAMPLE
CONCENTRATION
(ug/L)






MS
CONCENTRATION
(ug/L)






MS
%
REC #






QC
LIMITS
REC.
56-123
40-131
40-120
52-126
56-121
38-127
COMPOUND
gamma -BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4, 4 '-DDT
SPIKE
ADDED
(ug/L)






MSD
CONCENTRATION
(ug/L)



'N


MSD
%
REC #






%
RPD #






QC I
PRD
15
20
22
18
21
27
IMITS
REC.
56-123
40-131
40-120
52-126
56-121
38-127
 #  Column to be used to flag recovery and RPD values  with  an  asterisk

 *  Values outside of QC limits

 RPD:  	 out of 	 outside limits
 Spike Recovery:	 out of 	 outside limits

 COMMENTS:      •	
                                FORM III PEST-1
                                                                        OLM04.2

-------
                                      3F
          SOIL PESTICIDE  MATRIX  SPIKE/MATRIX  SPIKE  DUPLICATE  RECOVERY
Lab Name:

Lab Code:
Case No.:
Matrix Spike - EPA Sample No. :
  Contract:

SAS No.:
                  SDG No.:
COMPOUND
gamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4,4 ' -DDT
SPIKE
ADDED
(ug/Kg)






SAMPLE
CONCENTRATION
(ug/Kg)




. ;.

MS
CONCENTRATION
(ug/Kg)






MS
%
REC #






QC
LIMITS
REC.
46-127
35-130
34-132
31-134
42-139
23-134
COMPOUND
gamma - BHC (Lindane ) "
Heptachlor
Aldrin
Dieldrin
Endrin
4,4 ' -DDT
SPIKE
ADDED
(ug/Kg)






MSD
CONCENTRATION
'. (ug/Kg)






MSD
%
REC #






%
RPD #






QC L
RPD
50
31
43
38
45
50
IMITS
REC.
46-127
35-130
34-132
31-134
42-139
23-134
# Column to.be used to flag recovery and RPD values with an asterisk

* Values outside of QC limits

RPD:	out of 	 outside limits
Spike Recovery: 	 out of 	 outside limits
COMMENTS:
                                FORM III PEST-2
                                                                       OLM04.2

-------
                                      4A
                         VOLATILE METHOD  BLANK SUMMARY
                                          EPA SAMPLE NO.
Lab Name:

Lab Code:
                    Case No.:
                   Contract:

                   SAS No.:
                            SDG No.:
Lab File ID:
Date Analyzed:

GC Column: 	
                   Lab Sample ID:

                   Time Analyzed:
ID:
.(mm)  Heated Purge:  (Y/N)
Instrument ID:
       THIS METHOD BLANK APPLIES TO THE FOLLOWING SAMPLES, MS, AND MSD:
 COMMENTS:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























LAB
SAMPLE ID






























LAB
FILE ID






























TIME
ANALYZED






























 page
             of
                                  FORM IV VOA
                                                    OLM04.2

-------
                                      4B
                       SEMIVOLATILE METHOD BLANK SUMMARY
                                             EPA SAMPLE NO.
Lab Name -.

Lab Code:
Case No.:
Contract:

SAS No.:
                      SDG No.:
Lab File ID:
Instrument ID:
Matrix:  (soil/water)

Level:  (low/med) 	
                    Lab Sample ID:

                    Date Extracted:

                    Date Analyzed:

                    Time Analyzed:
       THIS METHOD BLANK APPLIES TO THE FOLLOWING SAMPLES, MS, AND MSD:
COMMENTS:
01
0-2
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.








j





















LAB
SAMPLE ID




..

























LAB
FILE ID






























DATE
ANALYZED






























page
            of
                                  FORM IV SV
                                                                       OLM04.2

-------
                                      4C
                        PESTICIDE METHOD BLANK SUMMARY
                                     EPA SAMPLE  NO.
Lab Name:

Lab Code:
                   Case No.:
              Contract:

              SAS No.:
                            SDG No.: .
Lab Sample ID:
               Lab  File  ID:
Matrix:  (soil/water)

Sulfur Cleanup:  (Y/N).

Date Analyzed  (1) : 	

Time Analyzed  (1) : 	

Instrument ID  (1): 	

GC Column  (1): 	
               Extraction:  (Type)

               Date  Extracted:
               Date  Analyzed (2):

               Time  Analyzed (2):

               Instrument ID (2):
ID:
.(mm)  GC Column  (2):
                                     ID:
(mm)
       THIS METHOD BLANK APPLIES  TO  THE  FOLLOWING SAMPLES,  MS,  AND MSD:
 COMMENTS:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
EPA
SAMPLE NO.


























LAB
SAMPLE ID


























DATE
ANALYZED 1


























DATE
ANALYZED 2


























 page
             of
                                  FORM IV PEST
                                               OLM04.2

-------
Lab Name:

Lab Code:
                                      5A
                 VOLATILE ORGANIC INSTRUMENT PERFORMANCE CHECK
                           BROMOFLUOROBENZENE  (BFB)
                     Contract:

                     SAS No.:
Case No.:
                                                          SDG No.:
Lab File ID:
Instrument ID:

GC Column: 	
                     BFB Injection Date:

                     BFB Injection Time:
    ID:
(mm)
m/e
50
75
95
96
173
174
175
176
177
ION ABUNDANCE CRITERIA
8.0 - 40.0% of mass 95
30.0 - 66.0% of mass 95
Base peak, 100% relative abundance
5.0 - 9.0% of mass 95
Less than 2.0% of mass 174
50.0 - 120.0% of mass 95
4.0 - 9.0 % of mass 174
93.0 - 101.0% of mass 174
5.0 - 9.0% of mass 176
% RELATIVE
ABUNDANCE




( )1

( )1
( )1
( 12
              1-Value  is  % mass  174
                          2-Value is % mass 176
 THIS CHECK APPLIES TO THE  FOLLOWING SAMPLES,  MS,  MSP,  BLANKS,'  AND STANDARDS:
            01
            02
            03
            04
            05
            06
           .07
            08
            09
            10
            11
            12
            13
            14
            15
            16
            17
            18
            19
            20
            21
            22
EPA
SAMPLE NO.






















LAB
SAMPLE ID






















LAB
FILE ID






















DATE
ANALYZED






















TIME
ANALYZED






















page
             of
                                   FORM V VGA
                                                                        OLM04.2

-------
                                      5B
               SEMIVOLATILE ORGANIC  INSTRUMENT PERFORMANCE  CHECK
                     DECAFLUOROTRIPHENYLPHOSPHINE (DFTPP)
Lab Name:

Lab Code:
Case No.:
                    Contract:

                    SAS No.:
SDG No.:
Lab File ID:
Instrument ID:
                    DFTPP Injection Date:

                    DFTPP Injection Time:
m/e
51
68
69
70
127
197
198
199
275
365
441
442
443
ION ABUNDANCE CRITERIA
30.0- 80.0% of mass 198
Less than 2.0% of mass 69
Mass 69 relative abundance
Less than 2.0% of mass 69
25.0 - 75.0% of mass 198
Less than 1.0% of mass 198
Base Peak, 100% relative abundance
5.0 to 9.0% of mass 198
10.0- 30.0% of mass 198
Greater than 0.75% of mass 198
Present, but less than mass 443
40.0 - 110.0% of mass 198
15.0 - 24.0% of mass 442
% RELATIVE
ABUNDANCE

( )1

( )1








( - )2
        1-Value is % mass 69
                       2-Value is % mass 442
 THIS  CHECK APPLIES  TO THE FOLLOWING SAMPLES,  MS,  MSP,  BLANKS, AND  STANDARDS;
            01
            02
            03
            04
            05
            06
            07
            08
            09
            10
            11
            12
            13
            14
            15
            16
            17
            18
            19
            20
            21
            22
EPA
SAMPLE NO.






















LAB
SAMPLE ID






















LAB
FILE ID






















DATE
ANALYZED






















TIME
ANALYZED






















 page
             of
                                    FORM V SV
                                                                        OLM04.2

-------
                                      6A
                  VOLATILE ORGANICS INITIAL CALIBRATION -DATA
Lab Name:

Lab Code:
Instrument ID:
                                Contract:
Case No.:
                    SAS No.:
Heated Purge:  (Y/N)

GC Column: 	
   ID:
 Calibration Date(s):

 Calibration Times:

	(mm)
                             SDG No.
LAB FILE ID: RRF10 =
RRF50 - RRF100 -
COMPOUND ,
Dichlorodif luoromethane
Chlorome thane
Vinyl Chloride *
Bromomethane *
Chloroethane
Trichlorof luoromethane
1, 1-Dichloroethene *
1, 1, 2-Trichloro- '
1,2 , 2-trif luoroethane
Acetone
Carbon Disulfide
Methyl Acetate
*1ethylene Chloride
trans-1, 2-Dichloroethene
Methyl tert -Butyl Ether
1,1-Dichloroethane *
cis-1, 2-Dichloroethene
2-Butanone
Chloroform *
1, 1, 1-Trichloroethane *
Cyclohexane .
Carbon Tetrachloride *
Benzene *
1,2-Dichloroethane *
Trichloroethene *
Methyl eye lohexane
RRF10



























RRF20

























RRF50

























RRF20
RRF200
RRF100

























_
-
RRF200


























RRF

























%
RSD


*
*


*







*


*
*

*
*
*
it

*Compounds with required minimum RRF and maximum %RSD values.
All other compounds must meet a minimum RRF of 0.010.
page
           of
                                 FORM VI VOA-1
                                                    OLM04.2

-------
                                      6B
                  VOLATILE ORGANICS INITIAL CALIBRATION DATA
Lab Name:

Lab Code:
Case No.:
Instrument ID:
Heated Purge:  (Y/N) 	

GC Column: 	 ID:
Contract:.

SAS No.:
          Calibration  Date(s):

          Calibration  Times:

         	(mm)
SDG No.
LAB FILE ID: RRF10 =
RRF50 - RRF100 -
COMPOUND
1 , 2 -Dichloropropane
Bromodichloromethane *
cis-1, 3-Dichloropropene *
4 -Methyl - 2 -pentanone
Toluene *
trans-1, 3-Dichloropropene *
1, 1, 2-Trichloroethane *
Tetrachloroethene *
2-Hexanone
Dibromochloromethane *
1, 2-Dibromoethane
Chlorobenzene *
Ethylbenzene *
Xylene (total) *
Styrene *
Bromoform *
Isopropylbenzene
1, 1, 2, 2-Tetrachloroethane *
1, 3-Dichlorobenzene *
1, 4-Dichlorobenzene *
1, 2-Dichlorobenzene *
1, 2-Dibromo-3-chloropropane
1,2,4-Trichlorobenzene *
Toluene -d8
Bromoflurobenzene *
1, 2-Dichloroethane-d4
RRF10



























RRF20
_
RRF200 -
RRF20


























RRF50


























RRF100






•



















RRF200



























RRF


























%
RSD

*
*

it
*
*
*

*

*
*
*
*
*

*
*
*
*

*

*

 *Compounds with required minimum RRF and maximum %RSD values.
 All other compounds must meet a minimum RRF of 0.010.
                                  FORM VI VGA-2
                                                     OLM04.2

-------
Lab Name:

Lab Code:
                       6C
 SEMIVOLATILE  ORGANICS INITIAL CALIBRATION DATA

	   Contract:	

    Case  No. :  	   SAS No. : 	 SDG No. :
Instrument ID:
          Calibration Date(s):

          Calibration Times:
LAB FILE ID: RRF20
RRF80 - RRF12

COMPOUND
Benzaldehyde
Phenol *
bis- (2-Chloroethyl) ether *
2-Chlorophenol *
2-Methylphenol *
2,2 ' -oxybis (1-Chloropropane)
Ac e t ophe none
4-Methylphenol *
N-Nitroso-di-n-propylamine *
Hexachloroethane *
Nitrobenzene *
Isophorone *
2-Nitrophenol *
2, 4-Dimethylphenol ' *
bis (2 -Chloroethoxy) methane *
2, 4-Dichlorophenol *
Naphthalene *
4-Chloroaniline
Hexachlorobutadiene
Caprolactam
4-Chloro-3-methylphenol *
2-Methylnaphthalene *
Hexachlorocyclopentadiene
2, 4, 6-Trichloropheriol *
2, 4, 5-Trichlorophenol *
1,1' -Biphenyl
2-Chloronaphthalene *
2 -Nitroaniline
Dimethylphthalate
2 , 6 -Dinitrotoluene *
Acenaphthylene' *
3 -Nitroaniline
Acenaphthene *
2 , 4 -Dinitrophenol
4 -Nitrophenol
Dibenzofuran *
	
0 =

RRF20







































RRF50







































RRF80




































RRF50
RRF160

RRF120




































	
=

RRF160



"



































RRF






"•"































%
RSD

*
*
*
*


*
*
*
*
*
•*
*
*
*
*



*
*

*
*

*


*
*

*


*
* Compounds with required minimum RRF and maximum %RSD values.
All other compounds must meet a minimum RRF of 0.010.
                                 FORM VI SV-1
                                                        OLM04.2

-------
Lab Name:

Lab Code:
                       6D
 SEMIVOLATILE  ORGANICS  INITIAL  CALIBRATION  DATA

	   Contract: 	

    Case No. :  	   SAS  No. :	 SDG No. :
Instrument ID:
          Calibration Date(s):

          Calibration Times:
LAB FILE ID: RRF20
=
RRF80 - RRF120 -
COMPOUND
2,4-Dinitrotoluene *
Diethylphthalate
Fluorene *
4-Chlorophenyl-phenylether *
4 -Nitroaniline
4, 6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine (1)
4-Bromophenyl-phenylether *
Hexachlorobenzene *
Atrazine
Pentachlorophe'nol' *
Phenanthrene *
Anthracene *
Carbazole
Di-n-butylphthalate
Fluoranthene *
Pyrene *
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Benzo (a) anthracene *
Chrysene *
bis(2-Ethylhexyl)phthalate
Di-n-octylphthalate
Benzo (b ) fluoranthene *
Benzo (k) fluoranthene *
Benzo (a)pyrene ' *
Indeno(l,2,3-cd)pyrene *
Dibenzo (a, h) anthracene *
Benzo (a,h, i)r>ervlene *
Nitrobenzene-d5 *
2-Fluorobiphenvl *
Terphenvl-dl4 *
Phenol -d5 *
2 - Fluorophenol *
2,4, 6-TribromoDhenol
2-Chlorot)henol-d4 *
1 . 2-Dichlorobpn7£?np-d4 *
RRF20





































RRF50





































RRF80





































RRF50
=
RRF160 -
RRF120





































RRF160






































RRF





































%
RSD
*

*
*



*
*

*
*
*


*
*


*
*


*
*
.*
it
*
*
*
*
*
*
*

*
*
 (1)  Cannot be separated from Diphenylamine
 *  Compounds with required minimum RRF and maximum %RSD values.
 All other compounds must meet a minimum RRF of 0.010.
                                  FORM VI SV-2
                                                        OLM04.2

-------
                                      6E
          PESTICIDE INITIAL CALIBRATION OF SINGLE COMPONENT ANALYTES
Lab Name:

Lab Code:
Instrument ID:

GC Column: 	;
Case No.:
                    Contract:

                    SAS No.:
      Level (x low):  low 	

      ID: 	(mm)  Date(s)  Analyzed:
	 SDG No.: 	

mid 	 high
COMPOUND
alpha-BHC
beta-BHC
delta-BHC
gamma-BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4,4' -DDE
Endrin
Endosulfan II
4,4' -ODD
Endosulfan sulfate
4,4' -DDT
Methbxychlor
Endrin ketone
Endrin aldehyde
alpha - Chlordane
gamma- Chlordane
Tetrachloro-m-xylene
Decachlorobiphenyl
RT
LOW






















OF STANDA
MID






















RDS
HIGH






















MEAN
RT






















RT W]
FROM






-















:NDOW
TO






















* Surrogate retention times are measured from Standard Mix A analyses.

Retention time windows are ± 0.05 minutes for all compounds that elute before
Heptachlor expoxide, ± 0.07 minutes for all other compounds, except ± 0.10
minutes for Decachlorobiphenyl.
                                FORM VI PEST-1
                                                    OLM04.2

-------
Lab Name:

Lab Code:
                            6F
PESTICIDE INITIAL CALIBRATION OF SINGLE COMPONENT ANALYTES


	  Contract: 	

                      	   SAS No.: 	
Instrument ID:

GC Column: 	
                   Case No.:
               Level  (x low): low 	

               ID: 	(mm) Date(s) Analyzed:
	 SDG No.:  	

mid 	 high
COMPOUND
alpha-BHC
beta-BHC
delta-BHC
gamma-BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4,4' -DDE
Endrin
Endosulfan II
4,4' -ODD
Endosulfan sulfate
4 ,4 ' -DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha- Chlordane
gamma - Chlordane
Tetrachloro-m-xylene
Decachlorobiphenyl
LOW






















CALIBRATIC
MID






















)N FACTORS
HIGH






















MEAN






















%RSD















:






 *  Surrogate  calibration factors are measured from Standard Mix A analyses.
                                 FORM VI PEST-2
                                                              OLM04.2

-------
                                      6G
           PESTICIDE INITIAL CALIBRATION OF MULTICOMPONENT ANALYTES
Lab Name:

Lab Code:
Case No.:
Contract:

SAS No.:
SDG No.:
Instrument ID:

GC Column: 	
      Date(s) Analyzed:

      ID: 	(mm)
COMPOUND
Toxaphene
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
AMOUNT
(ng)















PEAK1
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
. 5
1
2
3
4
5
RT








































RT WI
FROM








































NDOW
TO








































CALIBRATION
FACTOR








































lAt least 3 peaks for each column are required for identification of
multicomponent analytes.
                                FORM VI PEST-3
                                                    OLM04.2

-------
                                       6H
                     PESTICIDE ANALYTE RESOLUTION SUMMARY
Lab Name:

Lab Code:
                   Case No.
                                        Contract:

                                        SAS No.:
                          SDG  No.
GC Column  (1):
                          ID:
EPA Sample No.  (RESC##):

Date Analyzed  (1) : 	
.(mm)  Instrument  ID  (1):

	   Lab Sample  ID  (1):

	   Time Analyzed  (1):
01
02
03
04
05
06
07
08
09
ANALYTE









RT









RESOLUTION









mn
pie
aly


01
02
03
04
05
06
07
08
09
(2) : ID: (mm) Instr
No. (RESC##) : Lab S
zed (2) : Time

ANALYTE









ument ID (2) :
ample ID (2) :
Analyzed (2) :

RT













RESOLUTION
(%)























                                 FORM VI PEST-4
                                                                         OLM04.2

-------
                                       61
                     PERFORMANCE EVALUATION MIXTURE (PEM)
Lab Name:

Lab Code:
Case No.
      Contract:

      SAS No.:
SDG No.
GC Column  (1) :
       ID:
EPA Sample No.  (PEM##):

Date Analyzed  (1): 	
.(mm)   Instrument  ID (1) :

	    Lab  Sample  ID (1):

	    Time Analyzed (1):
01
02
03
04
05
06
07
08
ANALYTE





• - -


RT








RESOLUTION








mn
iple
aly


01
02
03
04
05
06
07
08
(2) : . ID: (mm) Instl
No. (PEM##) : Lab £
zed (2) : Time

ANALYTE








-ument ID (2) :
.ample ID (2) :
Analyzed (2) :

RT












RESOLUTION
(%)





















                                FORM VI PEST-5
                                                     OLM04.2

-------
                                      6J
                         INDIVIDUAL STANDARD MIXTURE A
Lab Name:

Lab Code:
                   Case No.
             Contract:

             SAS No.:
                          SDG No.
GC Column  (1) :
ID:
EPA Sample No.  (INDAM##):

Date Analyzed  (1): 	
.(mm)   Instrument  ID (1):

	    Lab Sample  ID (1):

	    Time Analyzed (1) :
01
02
03
04
05
06
07
08
09
10
11
ANALYTE








.


RT











RESOLUTION











mn (2) : % ID: (mm) Instrument ID (2) :
pie
aly
01
02
03
04
05
06
07
08
09
10
11
No. (INDAMtttf) : Lab £
zed (2) : Time

ANALYTE











ample ID (2) :
Analyzed (2) :
RT














RESOLUTION
(%)












                                 FORM VI PEST-6
                                              OLM04.2

-------
Lab Name:

Lab Code:
                    6K
      INDIVIDUAL STANDARD MIXTURE B

	   Contract:  	

 Case No. :  	   SAS No. :  	
                          SDG No.:
GC Column  (1) :
        ID:
EPA Sample No.  (INDBMtftf):

Date Analyzed  (1): 	
.(mm)   Instrument  ID  (1):

	    Lab  Sample  ID  (1):

	    Time Analyzed  (1):
01
02
03
04
05
06
07
08
09
10
11
12
13
ANALYTE













RT













RESOLUTION













GC Column  (2):
        ID:
EPA Sample No,  (INDBM##):

Date Analyzed  (2): 	
.(mm)   Instrument  ID (2):

	    Lab  Sample  ID (2):

	    Time Analyzed (2):
01
02
03
04
05
06
07
08
09
10
11
12
13
ANALYTE













RT













RESOLUTION













                                 FORM VI PEST-7
                                                      OLM04.2

-------
                                      7A
                    VOLATILE  CONTINUING CALIBRATION CHECK
Lab Name :
Lab Code :
Instrument ID:
Lab File ID:
Contract :
Case No.: SAS No.: SDG No.:
Calibration Date: Time:
Init. Calib. Date(s) :
EPA Sample No.(VSTD050##):

Heated Purge: (Y/N)	

GC Column: '	   ID:
     Init.  Calib.  Times:
(mm)
COMPOUND
Dichlorodif luoromethane
Chlorome thane
Vinyl Chloride
Bromome thane
Chloroethane
Trichlorof luoromethane
1, 1-Dichloroethene
1,1, 2-Trichloro-l, 2, 2-trif luoroethane
Acetone
Carbon Disulfide
Methyl Acetate
Methylene Chloride
trans- 1, 2-Dichloroethene
Methyl tert -Butyl Ether
1, 1-Dichloroethane
cis-1, 2-Dichloroethene
2-Butanone
Chloroform
1,1, 1-Trichloroethane
Cyclohexane
Carbon Tetrachloride
Benzene
1 , 2 -Dichloroethane
Trichloroethene
Methylcyclohexane

RRF

























RRF50

























MIN
RRF


0.100
0.100


0.100







0.200


0.200
0.100

0.100
0.500
0.100
0.300

%D

























MAX
%D


25.0
25.0


25.0







25.0


25.0
25.0

25.0
25.0
25.0
25.0

All  other  compounds must meet a minimum RRF of  0.010.
                                 FORM VII VOA-1
                                                                        OLM04.2

-------
Lab Name:

Lab Code:
Instrument ID:

Lab File ID:
                  7E
 VOLATILE CONTINUING CALIBRATION CHECK

	   Contract: 	

            	   SAS No.: 	
Case No.:
EPA Sample No.(VSTD050##):

Heated Purge:  (Y/N) ;	

GC Column:	ID:
SDG No.:
           Calibration Date:
                                            Time:
                   Init. Calib. Date(s):

                   Init. Calib. Times:
            .(mm)
COMPOUND
1,2-Dichloropropane
Bromodichloromethane
cis-1, 3-Dichloropropene
4'-Methyl - 2 -pentanone
Toluene
trans-1, 3-Dichloropropene
1, 1, 2-Trichloroethane
Tetrachloroethene
2-Hexanone
Dibromochloromethane
1 , 2 -Dibromoethane
Chlorobenzene
Ethylbenzene
Xylene (total)
Styrene
Bromof orm
Isopropylbenzene
1,1,2, 2-Tetrachloroethane
1 , 3 -Dichlorobenzene
1,4-Dichlorobenzene
1., 2 -Dichlorobenzene.
1, 2-Dibromo-3-chloropropane
1, 2,4-Trichlorpbenzene
Toluene -d8
Bromof lurobenzene
l,2-Dichloroethane-d4.

RRF


























RRF50


























MIN
RRF

0.200
0.200

0.400
0.100
0.100
0.200

0.100

0.500
0.100
0.300
0.300
0.100

0.300
0.600
0.500
0.400

0.200

0.200

%D



















.






MAX
%D
. „•
25.0
25.0

25.0
25.0
25.0
25.0

25.0

25.0
25:0
25.0
25.0
25.0

25.0
25.0
25.0
25.0

25.0

25.0

All other compounds must meet a minimum RRF of 0.010.
                                FORM VII VOA-2
                                                   OLM04.2

-------
Lab Name:

Lab Code:
                    7C
 SEMIVOLATILE CONTINUING CALIBRATION CHECK

	   Contract: 	

              	   SAS No.: 	
Case No.:
                                       SDG No.:
Instrument ID:

Lab File ID:
          Calibration Date:
                                                          Time:
EPA Sample No.(SSTD050##):

GC Column: 	 ID: 	
                     Init. Calib. Date(s) :.

                     Init. Calib. Times:
                (mm)
COMPOUND
Benzaldehyde
Phenol
bis- (2-Chloroethyl) ether
2-Chlorophenol
2 -Methylphenol
2,2' -oxybis (1-Chloropropane)
Acetophenone
4 -Methylphenol
N-Nitroso-di-n-propylamine
Hexachloroethane
Nit roberiz ene
Isophorone
2 -Nitrophenol
2 , 4 -Dimethylphenol
bis (2 -Chloroethoxy) methane
2 , 4 -Dichlorophenol
Naphthalene
4 - Chloroani 1 ine
Hexachlorobutadiene
Caprolactam
4 -Chloro- 3 -methylphenol
2 -Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6-Trichlorophenol
2,4, 5-Trichlorophenol
1,1' -Biphenyl
2-Chloronaphthalene
2 -Nitroaniline
Dimethylphthalate
2, 6-Dinitrotol'uene
Acenaphthylene
3 -Nitroaniline
Acenaphthene
2 , 4 -Dinitrophenol
4 -Nitrophenol
Dibenzofuran

RRF



































-
RRF50




































MIN
RRF

0.800
0.700
0.800
0.700


0.600
0.500
0.300
0.200
0.400
0.100
0.200
0.300
0.200
"0.700



0.200
0.400

0.200
0.200

0.800


0.200
0.900

0.900


0.800
%D




































MAX
%D

25.0
25.0
25.0
25.0


25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0



25.0
25.0

25.0
25.0

25.0


25.0
25.0

25.0


25.0
 All other compounds must meet a minimum RRF of 0.010.
                                  FORM VII  SV-1
                                                                        OLM04.2

-------
Lab Name:

Lab Code:
                    7D
 SEMIVOLATILE CONTINUING CALIBRATION CHECK

	   Contract:  	

              	   SAS No.: 	
                   Case No.:
Instrument ID:

Lab File ID:
            Calibration Date:
SDG No.: _

	 Time:
EPA Sample No.(SSTD050##): 	

GC Column: 	 ID:
                     Init. Calib. Date(s):

                     Init. Calib. Times:

                   (mm)
COMPOUND
2 , 4 -binitrotoluene
Diethylphthalate
Fluorene
4-Chlorophenyl-phenylether
4 -Nitroaniline
4, 6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine (1)
4 -Brpmophenyl-phenylether
Hexachlorobenzene
Atrazine
Pentachlorophenol
Phenanthrene
Anthracene
Carbazole
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Benzo (a) anthracene
Chrysene
bis (2-Ethylhexyl)phthalate
Di-n-octylphthalate
Benzo (b) f luoranthene
Benzo (k) f luoranthene
Benzo (a) pyrene
Indeno ( 1 , 2 , 3 - cd) pyrene
Dibenzo (a, h) anthracene
Benzo (g, h, i) perylene
Nitrobenzene -d5
2-Fluorobiphenyl
Terphenyl-dl4
Phenol -d5
2-Fluorophenol .
2,4, 6 -Tribromophenol
2-Chlorophenol-d4
1, 2-Dichlorobenzene-d4

RRF





































RRF50





































MIN
RRF
0.200

0.900
0.400



0.100
0.100

0.050
0.700
0.700


0.600
0.600


0.800
0.700


0.700
0.700
0.700
0.500
0.400
0.500
0.200
0.700
0.500
0.800
0.600

0.800
0.400
%D





































MAX
%D
25.0

25.0
25.0



25.0
25.0

25.0
25.0
25.0


25.0
25.0


25.0
25.0


25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0

25.0
25.0
(1) Cannot be separated from Diphenylamine
 All other compounds must meet a minimum RRF of 0.010.
                                 FORM VII  SV-2
                                                                       OLM04.2

-------
                                      7E
                  PESTICIDE CALIBRATION VERIFICATION SUMMARY
Lab Name :
Lab Code :
GC Column:

Case No. :
ID:
Contract :
SAS No. : SDG No. :
(mm) Init . Calib. Date(s):
EPA Sample No.  (PIBLK##):

Lab Sample ID  (PIBLK): 	

EPA Sample No.  (PEM##): _

Lab Sample ID  (PEM): 	
Date Analyzed:

Time Analyzed:

Date Analyzed:

Time Analyzed:
PEM
COMPOUND
alpha -BHC
beta-BHC
gamma-BHC (Lindane)
Endrin
4, 4 '-DDT .
Methoxychlor
RT






RT Wl
FROM






NDOW
TO






CALC
AMOUNT
(ng)






NOM
AMOUNT
(ng)






%D






4,4'-DDT % Breakdown (1)

Combined % Breakdown (1)
Endrin % breakdown  (1):.
                                FORM VII PEST-1
                                 OLM04.2

-------
                                      7F
                  PESTICIDE CALIBRATION VERIFICATION SUMMARY
Lab Name :
Lab Code :
GC Column:

Case No. :
ID:
Contract :
SAS No. :
(mm) Init. Calib.

SDG No . :
Date(s) :
EPA Sample No.  (PIBLK##):

Lab Sample ID  (PIBLK): 	
EPA Sample No.  (INDAM##):

Lab Sample ID   (INDA): 	
Date Analyzed:

Time Analyzed:

Date Analyzed:

Time Analyzed:
INDIVIDUAL MIX A
COMPOUND
alpna-BHC
gamma -BHC (Lindane)
Heptachlor
Endosulfan I
Dieldrin
Endrin
4,4' -ODD
4,4' -DDT
Methoxychlor
Tetrachloro-m-xylene
Decachlorobiphenyl
RT











RT W]
FROM











ENDOW
TO











CALC
AMOUNT
(ng)











NOM
AMOUNT
(ng)











%D











EPA Sample No.  (INDBM##)

Lab Sample ID   (INDB): _
Date Analyzed:

Time Analyzed:
INDIVIDUAL MIX B
COMPOUND
tieta-BHC
delta-BHC
Aldrih
Heptachlor epoxide
4, 4 '-DDE
Endosulfan II
Endosulfan sulfate
Endrin ketone
Endrin aldehyde
alpha- Chlordane
gamma- Chlordane
Tetrachloro-m-xylene
Decachlorobiphenyl
RT













RT W]
FROM













CNDOW
TO













CALC
AMOUNT
(ng)













NOM
AMOUNT
(ng)













%D













                                FORM VII PEST-2
                                OLM04.2

-------
Lab Name:

Lab Code:
                       8A
 VOLATILE INTERNAL STANDARD AREA AND RT SUMMARY

	   Contract:  	

    Case No. :  	   SAS No. : 	
SDG No.:
EPA Sample No.(VSTD050##):

Lab File ID  (Standard): 	

Instrument ID: 	

GC Column:	ID: 	
                        Date Analyzed:

                        Time Analyzed:
                        Heated Purge:  (Y/N).
                 .(mm)

12 HOUR STD
UPPER LIMIT
LOWER LIMIT
EPA SAMPLE












'









IS1 (BCM)
AREA #


























RT #


























IS2 (DFB)
AREA #


























RT #


























IS3 (CBZ)
AREA #


























RT #


























 01
 02
 03
 04
 05
 06
 07
 08
 09
 10
 11
 12
 13
 14
 15
 16
 17
 18
 19
 20
 21
 22
 IS1  (BCM)  = Bromochloromethane
 IS2  (DFB)  = 1,4-Difluorobenzene
 IS3  (CBZ)  = Chlorobenzene-d5

 AREA UPPER LIMIT = + 100% of internal standard area
 AREA LOWER LIMIT = - 50%  of internal standard area
 RT UPPER LIMIT = + 0.50 minutes of internal standard RT
 RT LOWER LIMIT = - 0.50 minutes of internal standard RT

 # Column used to flag values outside QC limits with an asterisk.
 * Values outside of QC limits
 page_
             of
                                 FORM VIII VOA
                                                         OLM04.2

-------
Lab Name:

Lab Code:
                         SB
 SEMIVOLATILE INTERNAL STANDARD AREA AND RT SUMMARY

	   Contract:  	

                  	   SAS No.:  	
Case No.:
EPA Sample No. (SSTD050M) :

Lab File ID (Standard): 	

Instrument ID: 	
SDG No.:
                          Date Analyzed:

                          Time Analyzed:

                          GC Column:  	
                                            ID:
              (mm)

12 HOUR STD
UPPER LIMIT
LOWER LIMIT
EPA SAMPLE
NO.









T*
-."'
* •

"








ISl (DCB)
AREA '#


























RT #


























IS2 (NPT)
AREA #


























RT #


























IS3 (ANT)
AREA #


























RT #
-


•o-






















01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
ISl  (DCB) = l,4-Dichlorobenzene-d4
IS2  (NPT) = Naphthalene-d8
IS3  (ANT) = Acenaphthene-dlO

AREA UPPER LIMIT = + 100% of  internal standard area
AREA LOWER LIMIT = - 50%  of  internal standard area
RT UPPER LIMIT = + 0.50 minutes of internal standard RT
RT LOWER LIMIT = - 0.50 minutes of internal standard RT

# Column used to flag values  outside QC limits with an asterisk.
* Values outside of QC limits
page.
            of
                                 FORM VIII SV-1
                                                          OLM04.2

-------
Lab Name:

Lab Code:
                         8C
 SEMIVOLATILE INTERNAL STANDARD AREA AND  RT SUMMARY

    '	   Contract:  	

	 Case No. :  	.    SAS No. : 	 SDG No. :
EPA Sample No.(SSTD050##):

Lab File ID  (Standard): 	

Instrument ID: 	
                          Date Analyzed:

                          Time Analyzed:

                          GC Column:  	
ID:
.(mm)

12 HOUR STD
UPPER LIMIT -
LOWER LIMIT
EPA SAMPLE
NO.






















IS4 (PHN)
AREA #


























RT #



X






















IS5 (CRY)
AREA #


























RT #


























IS6 (PRY)
AREA #


























RT #


























 01
 02
 03
 04
 05
 06
 07
 08
 09
 10
 11
 12
 13
 14
 15
 16
 17
 18
 19
 20
 21
 22
 IS4  (PHN)  = Phenanthrene-dlO
 IS5  (CRY)  = Chrysene-dl2
 IS6  (PRY)  = Perylene-dl2

 AREA UPPER LIMIT = + 100% of internal standard area
 AREA LOWER LIMIT = - 50%  of internal standard area
 RT UPPER LIMIT = + 0.50 minutes of internal standard RT
 RT LOWER LIMIT = - 0.50 minutes of internal standard RT

 # Column used to flag values outside QC limits with an asterisk.
 * Values outside of QC limits
 page_
             of
                                 FORM VIII SV-2
                                                          OLM04.2

-------
Lab Name:

Lab Code:

GC Column:
              8D
 PESTICIDE ANALYTICAL SEQUENCE

	   Contract: 	

       	   SAS No.:  	
Case No. :  	            	

ID: 	(mm)  Init. Calib. Date(s):
SDG No.:
Instrument ID:
THE ANALYTICAL SEQUENCE OF PERFORMANCE EVALUATION MIXTURES,  BLANKS,  SAMPLES,
AND STANDARDS IS GIVEN BELOW:
MEAN SURROGATE RT FROM INITIAL CALIBRATION
TCX : • DCB :
EPA
SAMPLE NO .
































LAB
SAMPLE ID
































DATE
ANALYZED
































TIME
ANALYZED





-



























TCX
. RT #
































DCB
RT #
































 01
 02
 03
 04
 05
 06
 07
 08
 09
 10
 11
 12
 13
 14
 15
 16
 17
 18
 19
 20
 21
 22
 23
 24
 25
 26
 27
 28
 29
 30
 31
 32
TCX = Tetrachloro-m-xylene
DCB = Decachlorobiphenyl
             QC LIMITS
             (± 0.05 MINUTES)
             (± 0.10 MINUTES)
# Column used  to  flag  retention time  values  with 'an asterisk.
* Values outside  of  QC limits.
page
            of
                                 FORM VIII  PEST
                                               OLM04.2

-------
                                      9A
                      PESTICIDE FLORISIL CARTRIDGE CHECK
Lab Name:

Lab Code:
                   Case No.:
Florisil Cartridge Lot Number: 	

GC Column  (1) : 	 ID:	(mm)
Contract:  	

SAS No. : 	 SDG No. :  	

Date of Analysis: 	

GC Column(2) :  	 ID:
(mm)
COMPOUND
alpna-BHC
gamma-BHC (Lindane)
Heptachlor
Endosulfan I
Dieldrin
Endrin
4,4' -ODD
4,4 ' -DDT
Methoxychlor
Tetrachloro-m-xylene
Decachlorobiphenyl • ••
2,4. 5-Trichloror>henol
SPIKE
ADDED
(ng)












SPIKE
RECOVERED
(ng)












%
REC #












QC
LIMITS
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
<5
# Column  to be  used to flag recovery with an asterisk.
* Values  outside  of QC limits.

   THIS CARTRIDGE LOT APPLIES TO THE FOLLOWING SAMPLES, BLANKS, MS, AND MSD:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
EPA
SAMPLE NO.























LAB
SAMPLE ID























DATE
ANALYZED 1























DATE
ANALYZED 2























 page
             of
                                 FORM IX PEST-1
                                                                        OLM04.2

-------
                                      9B
                    PESTICIDE GPC CALIBRATION VERIFICATION
Lab Name:

Lab Code:
GPC Column:
Case No.:
                    Contract:

                    SAS No.:
                         SDG No.:
GC Column(1):
     ID:
	   Calibration Verification Date: 	

.(mm)  GC Column(2) : 	 ID: 	(mm)
COMPOUND
gamma-BHC (Liridane)
HeTDtachlor
Aldrin
Dieldrin
Endrin
4.4' -DDT
SPIKE
ADDED
(ng)






SPIKE
RECOVERED
(ng)






%
REC #






QC
LIMITS
80-110
80-110
80-110
80-110
80-110
80-110
# Column to be used to flag recovery with an asterisk.
* Values outside of QC limits.

 THIS GPC CALIBRATION VERIFICATION APPLIES TO THE FOLLOWING SAMPLES, BLANKS,
MS, AND MSD:
01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
. EPA
SAMPLE NO.


























LAB
SAMPLE ID


























DATE
ANALYZED 1


























DATE '
ANALYZED 2


























page
            of
                                 FORM  IX  PEST-2
                                                                        OLM04.2

-------
                                      10A
                       PESTICIDE IDENTIFICATION  SUMMARY
                         FOR SINGLE COMPONENT ANALYTES
Lab Name:

Lab Code:
Case No.:
Lab Sample ID:
Instrument ID  (1):

GC Column:(1): 	
      ID:
      Contract:

      SAS No.:
.(mm)  GC Column:(2):
                                                                 EPA SAMPLE NO.
                                       SDG No.:
                    Date(s) Analyzed:

                    Instrument ID  (2):
                                               ID:
.(mm)
ANALYTE









COL
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
RT
















RT WI
FROM
















NDOW
TO
















CONCENTRATION
















%D











 page
             of
                                  FORM X PEST-1
                                                     OLM04.2

-------
Lab Name:

Lab Code:
                                      10B
                       PESTICIDE IDENTIFICATION SUMMARY
                          FOR MULTICOMPONENT ANALYTES
                    Contract:

                    SAS No.:
                                             EPA SAMPLE NO.
Case No.:
                                                           SDG No.:
Lab Sample ID:
Instrument ID  (1):

GC Column:(1):  	
                    Date(s) Analyzed:

                    Instrument ID  (2):
      ID:
.(mm)  GC Column:(2):
                                               ID:
(mm)
ANALYTE

COLUMN 1
COLUMN 2

COLUMN 1
COLUMN 2

COLUMN 1
COLUMN 2
PEAK
1
2
3
4
5 '
1.
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
• 5
RT



































RT W]
FROM






.,




























NDOW
TO



































CONCENTRATION





J-





























MEAN
CONCENTRATION






%D




At least 3 peaks for each column  are  required for  identification of
multicomponent analytes.
page
            of
                                 FORM X PEST-2
                                                                        OLM04.2

-------
                                       SAMPLE LOG-IN SHEET
Lab Name
Received By (Print Name)
Page 	 of 	 1
c
Log- in Date
Received By (Signature)
Case Number
Remarks :
1. Custody Seal(s)
2 . Custody Seal Nos .
3 . Chain of Custody
Records
4 . Traffic Reports
or Packing Lists
5. Airbill
6. Airbill No.
7 . Sample .Tags
Sample Tag
Numbers
8 . Sample Condition
9. Cooler
Temperature
10. Does information
on custody
records, traffic
reports, and
sample tags
agree?
11. Date Received at
Lab
12. Time Received

Present /Absent*
Intact /Broken


Present /Absent*
Present /Absent*
Airbill/Sticker
Present /Absent*

Present /Absent*
Listed/Not
Listed on Chain-
of -Custody
Intact/Broken* /
Leaking
Yes/No*

Sample Transfer
Fraction
Area #
By
On
Fraction
Area #
By
On
Sample Delivery Group No.

EPA Sample
#

















Corresponding
Sample Tag
#

















Assigned Lab #

















SAS Number

Remarks :
Condition of
Sample
Shipment , etc .






I










* Contact SMO and attach record of resolution
Reviewed By
Date
Logbook No. f
Logbook Page No.
                                            FORM DC-1
OLM04.2

-------
                  ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET
 LABORATORY NAME,
 CITY/STATE 	
 CASE NO. 	 SDG NO.	 SDG NOS. TO  FOLLOW_
 	SAS NO.	
 CONTRACT NO.
 SOW NO. 	
All documents delivered in the Complete SDG File must be original documents where
possible.
                                                              PAGE NOs             CHECK
                                                            FROM       TO       LAB      EPA
 1. Inventory Sheet (Form DC-2)  (Do not Number)           	  	  	  	
 2. SDG Case Narrative                                   	  	  	  	
 3. SDG Cover Sheet/Traffic Report                       	  	  	  	
 4. Volatiles Data
    a.  QC Summary
        System Monitoring Compound Summary  (Form II        .	  	  	  	
        Matrix Spike/Matrix Spike Duplicate Summary
              (Form III VOA)                              	  	  	  	
        Method Blank Summary  (Form IV VOA)               	     '      	  	
        GC/MS Instrument Performance Check  (Form V VOA)  	  	  	  	
        Internal Standard Area and RT Summary
              (Form VIII VOA)                             	  	  	  	
    b.  Sample Data
        TCL Results  -  (Form I VOA-1, VOA-2)
        Tentatively  Identified Compounds  (Form I VOA-
        Reconstructed  total ion chromatograms  (RIC) for
             each sample
        For each sample:
           Raw Spectra and background-subtracted mass
             spectra of target compounds identified
           Quantitation reports
           Mass Spectra of all reported TICs with three
             best library matches

    c.  Standards Data  (All Instruments)
        Initial Calibration Data  (Form VI VOA-1, VOA-2)
        RICs and Quan  Reports for all Standards
        Continuing Calibration Data
            (Form VII VOA-1, VOA-2)
        RICs and Quantitation Reports for all Standards

    d.  Raw QC Data
        BFB
        Blank Data
        Martix Spike/Matrix Spike Duplicate Data
                                    FORM DC-2-1                             OLM04.2

-------
             ORGANICS COMPLETE SDG FILE (CSF)  INVENTORY SHEET (cont.)
 CASE  NO. 	 SDG NO.  	 SDG NOS.  TO FOLLOW .
 	SAS NO.    '	
                                                             PAGE NOs            CHECK
                                                          FROM        TO       LAB      EPA
5.  Semivolatlies Data
   a.   QC Summary
       Surrogate Percent Recovery Summary (Form II SV)
       MS/MSD Summary (Form III SV)
       Method Blank Summary (Form IV SV)
       GC/MS Instrument Performance Check (Form V SV)
       Inte'rnal Standard Area and RT Summary
            (Form VIII SV)
   b.  Sample Data
       TCL Results -  (Form I SV-1, SV-2)
       Tentatively Identified Compounds  (Form I SV-
       Reconstructed total ion chromatograms (RIC) for
            each sample
       For each sample:
          Raw Spectra and background-subtracted mass
            spectra of target compounds
          Quantitation reports
          Mass Spectra of TICs with three best library
            matches
          GPC chromatograms  (if GPC is required)

   c.  Standards Data  (All Instruments)
       Initial Calibration Data  (Form VI SV-1, SV-2)
       RICs and Quan Reports for  all Standards
       Continuing Calibration Data  (Form VII SV-1, SV-
       RICs and Quantitation Reports for all Standards

   d.  Raw QC Data
       DFTPP
        /
       Blank Data
       Matrix Spike/Matrix Spike  Duplicate Data

   e.  Raw GPC Data

 6. Pesticides Data
   a.  QC Summary
       Surrogate Percent  Recovery Summary  (Form  II
       MS/MSD Duplicate  Summary  (Form  III  PEST)
       Method Blank  Summary  (Form IV PEST)
                                   FORM DC-2-2                             OLM04 . 2

-------
             ORGANICS COMPLETE SDG FILE (CSF)  INVENTORY SHEET (cont.)
 CASE  NO. 	 SDG NO.  	 SDG NOS.  TO FOLLOW .
                                 	 SAS NO. 	
                                                             PAGE NOs            CHECK
                                                          FROM        TO      LAB     EPA
6.  Pesticides Data (Cont.)
   b.   Sample Data                                      	  	
       TCL Results - Organic Analysis Data Sheet
            (Form I PEST)
       Chromatograms  (Primary Column)
       Chromatograms from second GC column
       GC Integration report or data system printout
       Manual work sheets
       For pesticides/Aroclors by GC/MS,
          Copies of raw spectra and copies of
          background-subtracted mass spectra of target
          compounds  (samples & standards)
   c.   Standards Data
       Initial Calibration of Single Component
             (Form VI PEST-1 and PEST-2)
       Initial Calibration of Multicomponent Analytes
             (Form VI PEST-3)
       Analyte Resolution Summary  (Form VI PEST-4)
       Performance Evaluation Mixture  (Form VI PEST-5)
     •  Individual Standard Mixture A (FORM VI PEST-6)
       Individual Standard Mixture B (FORM VI PEST-7)
       Calibration Verification Summary
             (Form VII PEST-1)
       Calibration Verification Summary
             (Form VII PEST-2)
       Analytical Sequence  (Form VIII PEST)
       Florisil Cartridge Check  (Form IX PEST-1)
       Pesticide GPC Calibration  (Form IX PEST-2)
       Pesticide Identification Summary for Single
            Component Analytes  (Form X PEST-1)
       Pesticide Identification Summary for
            Multicomponent Analytes  (Form X PEST-2)
       Chromatograms and data system printouts
          A printout of retention times and
          corresponding peak areas or peak heights

   d.   Raw QC Data
       Blank Data
       Matrix Spke/Matrix Spike Duplicate Data
                                  FORM  DC-2-3                             OLM04 . 2

-------
             ORGANICS COMPLETE SDG FILE (CSF)  INVENTORY SHEET (cont.)
 CASE NO.
                     SDG NO.
SDG NOS.  TO FOLLOW.
SAS NO. 	
6.  Pesticides Data (Cont.)
   e.   Raw GPC Data
                                                            PAGE NOs
                                                         FROM        TO
                                         CHECK
                                      LAB     EPA
   f.  Raw Florisil Data
7.  Miscellaneous Data
      Original preparation and analysis forms or
        of preparation and analysis logbook pages
      Internal sample and sample extract transfer
        chain-of-custody records
      Screening records
      All instrument output, including strip charts
        from screening activities  (describe or list)
8. EPA Shipping/Receiving Documents
     Airbills  (No. of shipments	)
     Chain-of-Custody Records
     Sample Tags
     Sample Log-in Sheet  (Lab & DC1)
     Miscellaneous Shipping/Receiving  Records
        (describe or list)
 9. Internal Lab Sample Transfer Records and Tracking
   Sheets  (describe or list)
 10. Other Records (describe or list)
       Telephone Communication Log
 11. Comments;
                                   FORM DC-2-4
                                                                           OLM04 . 2

-------
             ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET (cont.)
CASE NO.  	 SDG NO. 	 SDG NOS. TO FOLLOW

                                  	 SAS NO. 	
Completed by:   _____^^^_^^_^^_^^_                 	       _^^__^^^^^
(CLP Lab)              (Signature)(Printed Name/Title)        (Date)


Verified by:    		  _^_^_^	^_^_^_  	^_^.
(CLP Lab)              (Signature)          (Printed Name/Title)        (Date)

Audited by:
(EPA)                  (Signature)          (Printed Name/Title)        (Date)
                                  FORM  DC-2-5                            OLM04.2

-------
                                  EXHIBIT C
                           TARGET COMPOUND LIST AND
                    CONTRACT REQUIRED QUANTITATION LIMITS
NOTE: Specific quantitation limits are highly matrix-dependent.  The
      guantitation limits listed herein are provided for guidance and may not
      always be achievable.

      All CRQLs are rounded to two significant figures.

      The CRQL values listed on the following pages are based on the analysis
      of samples according to the specifications given in Exhibit D.

      For soil samples, the moisture content of the samples must be used to
      adjust the CRQL values appropriately.
                                      C-l                                OLM04.2

-------
Exhibit C - Target Compound List and Contract Required Quantitation Limits

                               Table  of  Contents

Section            .                                                       Page

1.0   VOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED QUANTITATION
      LIMITS	3

2.0   SEMIVOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED QUANTITATION
      LIMITS	,	5

3.0   PESTICIDES/AROCLORS TARGET COMPOUND LIST AND CONTRACT REQUIRED
      QUANTITATION LIMITS 	 	  8
                                      C-2                                OLM04.2

-------
                                                         Exhibit C -- Section 1
                                                                Volatiles  (VOA)

1.0   VOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED QUANTITATION LIMITS
Ouantitation Limits

1.
2.
3.
4.
5.
6.
7.
8.

9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
31.

32.
33.
34.
35.
Volatiles
Di chl orodi fl uorome thane
Chlorome thane
Vinyl Chloride
Bromomethane
Chloroethane
Tri chl orofl uorome thane
1, 1-Dichloroethene
1, 1,2-Trichloro-
1,2,2-trifl uoroe thane
Acetone
Carbon Disulfide
Methyl Acetate
Methylene Chloride
trans-1, 2-Dichloroethene
Methyl tert -Butyl Ether
1 , 1-Dichloroe thane
cis- 1, 2-Dichloroethene
2-Butanone
Chloroform
1,1, 1-Trichloroethane
Cyclohexane
Carbon Tetrachloride
Benzene
1 , 2 -Dichloroethane
Trichloroethene
Me thyl cycl ohexane
1 , 2 -Dichloropropane
Bromodichloromethane
cis-1 , 3-Dichloropropene
4 -Methyl -2 -pentanone
Toluene
trans -1,3-
Dichloropropene
1,1, 2-Trichloroethane
Tetrachloroethene
2-Hexanone
Dibromochloromethane
CAS Number
75-71-8
74-87-3
75-01-4
74-83-9
75-00-3
75-69-4
75-35-4
76-13-1

67-64-1
75-15-0
79-20-5
75-09-2
156-60-5
1634-04-4
75-34-3
156-59-2
78-93-3
67-66-3
71-55-6
110-82-7
56-23-5
71-43-2
107-06-2
79-01-6
108-87-2
78-87-5
75-27-4
10061-01-5
108-10-1
108-88-3
10061-02-6

79-00-5
127-18-4
591-78-6
124-48-1
Water
ua/L
10
10
10
10
10
10
10
10

10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10

10
10
10
10
Low
Soil
uq/Kq
10
10
10
10
10
10
10
10

10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10

10
10
10
10
Med.
Soil
uq/Kq
1200
1200
1200
1200
1200
1200
1200
1200

1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200

1200
1200
1.200
1200
On
Column
(nq)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)

(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(5.0)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)

(50)
(50)
(50)
(50)
                                       C-3
OLM04.2

-------
Exhibit C -- Section 1
Volatiles (VGA)
 1.0   VOLATILES  TARGET COMPOUND LIST AND CONTRACT REQUIRED QUANTITATION LIMITS
       (Con't)
Ouantitation Limits

36.
37.
38.
39.
40.
41.
42.
43.

44.
45.
46.
47.
48.
Volatiles
1 , 2 -Dibromoethane
Chlorobenzene
Ethylbenzene
Xylenes (total)
Styrene
Bromoform
Isopropylbenzene
1,1,2,2-
Tetrachloroethane
1,3-Di chlorobenzene
1,4-Di chlorobenzene
1, 2-Dichlorobenzene
'. 1, 2-Dibromo-3-chloropropane
1,2,4- Trichlorobenzene
CAS Number
106-93-4
108-90-7
100-41-4
1330-20-7
100-42-5
75-25-2
98-82-8
79-34-5

541-73-1
106-46-7
95-50-1
96-12-8
120-82-1
Water
UQ/L
10
10
10
10
10
10
10
10

10
10
10
10
10
Low
Soil
uq/Kq
10
10
10
10
10
10
10
10

10
10
10
10
10
Med.
Soil
uq/Kq
1200
1200
1200
1200
1200
1200
1200
1200

1200
1200
1200
1200
1200
On
Column
(nq)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)

(50)
(50)
(50)
(50)
(50)
                                       C-4
                                                                         OLM04.2

-------
2.0
                                                   Exhibit C -- Section 2
                                                     Semivolatiles .(SVGA)

SEMIVOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED QUANTITATION
LIMITS
Ouantitation Limits

49.
' 50.
. 51.

52.
53.
54.

55.
56.
57.

58.
59.
60.
61.
62.
63.

64.
65.
66.
67.
68.
69.

70.
71.

72.
73.
Semivolatiles CAS
Benzaldehyde
Phenol
bis- (2-Chloroethyl)
ether
2 - Chi or opheno 1
2-Methylphenol
2,2 ' -oxybis (1-
Chloropropane ) 1
Acetophenone
4 -Methylphenol
N-Nitroso-di-n
propylatnine
Hexachloroethane
Nitrobenzene
Isophorone
2 -Nitrophenol
2 , 4 -Dimethylphenol
bis (2-Chloroethoxy)
methane
2 , 4 -Dichlorophenol
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
Caprolactam
4-Chloro-3-
methylphenol
2 -Methylnaphthalene
Hexachlorocyc lo -
pentadiene
2,4, 6 -Trichlorophenol
2 , 4 , 5-Trichlorophenol
Number
100-52-7
108-95-2
111-44-4

95-57-8
95-48-7
108-60-1

98-86-2
106-44-5
621-64-7

67-72-1
98-95-3
78-59-1
88-75-5
105-67-9
111-91-1

120-83-2
91-20-3
106-47-8
87-68-3
105-60-2
59-50-7

91-57-6
77-47-4

88-06-2
95-95-4
Water
UQ/-L
10
10
10

10
10
10

10
10
10

10
10
10
10
10
10

10
10
10
10
10
10

10
10

10
25
Low
Soil
uq/Kq
330
330
330

330
330
330

330
330
330

330
330
330
330
330
330

330
330
330
330
330
330

330
330

330
830
Med.
Soil
uq/Kq
10000
10000
10000

10000
10000
10000

10000
10000
10000

10000
10000
10000
10000
10000
10000

10000
10000
10000
10000
10000
10000

10000
10000

10000
25000
On
Column
(nq)
(20)
(20)
(20)

(20)
(20)
(20)

(20)
(20)
(20)

(20)
(20)
(20)
(20)
(20)
(20)

(20)
(20)
(20)
(20)
(20)
(20)

(20)
(20)

(20)
(50)
      1 Previously known by the name bis(2-Chloroisopropyl)ether.

                                       C-5
                                                                   OLM04.2

-------
Exhibit C -- Section 2
Semivolatiles  (SVGA)
2.0    SEMIVOLATILES  TARGET  COMPOUND LIST AND CONTRACT REQUIRED QUANTITATION
       LIMITS  (Con't)
Ouantitation Limit

74.
75.
76.
77.
78.
79.
80.
81.
82.
83 .
84.
85.
86.
87.
88.

89.
90.

91.

92.

93.
94.
95.
96.
97.
98.
99.
100.
101.
102.
103.

104.
105.
Semivolatiles
1,1' -Biphenyl
2-Chloronaphthalene
2 -Nitroaniline
Dimethylphthalate
2 , 6-Dinitrotoluene
Acenaphthylene
3 -Nitroaniline
Acenaphthene
2 , 4 -Dinitrophenol
4 -Nitrophenol
Dibenzofuran
2 , 4-Dinitrotoluene
Diethylphthalate
Fluorene
4-Chlorophenyl-
phenyl ether
4 -Nitroaniline
4, 6-Dinitro-2-
methylphenol
N-Nitroso
diphenylamine
4 - Bromophenyl -
phenylether
Hexachlorobenzene
Atrazine
Pehtachlorophenol
Phenanthrene
Anthracene
Carbazole
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3'-
Dichlorobenz idine
Benzo (a) anthracene
Chrysene
CAS Number
92-52-4
91-58-7
88-74-4
131-11-3
606-20-2
208-96-8
99-09-2
83-32-9
51-28-5
100-02-7
132-64-9
121-14-2
84-66-2
86-73-7
7005-72-3

100-01-6
534-52-1

86-30-6

101-55-3

118-74-1
1912-24-9
87-86-5
85-01-8
120-12-7
86-74-8
84-74-2
206-44-0
129-00-0
85-68-7
91-94-1

56-55-3
218-01-9
Water
UQ/L
10
10
25
10
10
10
25
10
25
25
10
10
10
10
10

25
25

10

10

10
10
25
10
10
10
10
10
10
10
10

10
10
Low
Soil
ua/Ka
330
330
830
330
330
330
830
330
830
830
330
330
330
330
330

830
830

330

330

330
330
830
330
330
330
330
330
330
330
330

330
330
Med.
Soil
UQ/Kq
10000
10000
25000
10000
10000
10000
25000
10000
25000
25000
10000
10000
10000
10000
10000

25000
25000

10000

10000

10000 -
10000
25000
10000
10000
10000
10000
10000
10000
10000
10000

10000
10000
On
Column
(net)
(20)
(20)
(50)
(20)
(20)
(20)
(50)
(20)
(50)
(50)
(20)
(20)
(20)
. (20)
(20)

(50)
(50)

(20)

(20)

(20)
(20)
(50)
(20)
(20)
(20)
(20)
(20)
(20)
(20)
(20)

(20)
(20)
                                       C-6
                                                                         OLM04.2

-------
2.0
                                                    Exhibit C -- Section 2
                                                      Semivolatiles  (SVOA)

SEMIVOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED QUANTITATION
LIMITS (Con't)
                                                   Ouantitation Limits
Semivolatiles
Water
CAS Number uq/L
Low
Soil
uq/Kq
Med.
Soil
uq/Kq
On
Column
(nor)
 106.   bis(2-Ethylhexyl)         117-81-7     10      330     10000    (20)
       phthalate
 107.   Di-n-octylphthalate       117-84-0     10      330     10000    (20)
 108.   Benzo(b)fluoranthene      205-99-2     10      330     10000    (20)
 109.   Benzo(k)fluoranthene      207-08-9     10      330     10000    (20)
 110.   Benzo(a)pyrene  -           50-32-8     10      330     10000    (20)
 111..  Indeno(l,2,3-cd)-         193-39-5     10      330     10000    (20)
       pyrene
 112.   Dibenzo(a.h)-              53-70-3     10      330     10000    (20)
       anthracene
 113.   Benzo(g,h,i)perylene      191-24-2     10      330     10000    (20)
                                       C-7
                                                                         OLM04.2

-------
     EXHIBIT  D
ANALYTICAL METHODS
   FOR VOLATILES
      D-l/VOA                            OLM04.2

-------
Exhibit C -- Section 3
Pesticides/Aroclors  (PEST/ARO)

3.0     PESTICIDES/AROCLORS  TARGET  COMPOUND  LIST AND  CONTRACT REQUIRED
        QUANTITATION  LIMITS2
Ouantitation


Pesticides/Aroclors
114.
115.
116.
117.
118.
119.
120.
121.
122.
123.
124.
125.
126.
127.
128.
129.
130.
131.
132.
133.
134.
135.
136.
137.
138.
139.
140.
141.
alpha-BHC
beta-BHC
delta-BHC
gamma-BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide3
Endosulfan I
Dieldrin
4,4' -DDE
Endrin
Endosulfan II
4,4' -ODD
Endosulfan sulfate
4,4' -DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
a Ipha - Chi or dane
gamma- Chlordane
Toxaphene
Aroclor-1016
Aroclor-1221
Aroclor-1232
Aroclor-1242
Aroclor-1248
Aroclor-1254
Aroclor-1260

CAS Number
319-84-6
319-85-7
319-86-8
58-89-9
76-44-8
309-00-2
1024-57-3
959-98-8
60-57-1
72-55-9
72-20-8
33213-65-9
72-54-8
1031-07-8
50-29-3
72-43-5
53494-70-5
7421-93-4
5103-71-9
5103-74-2
8001-35-2
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
Water
ua/L
0.050
0.050
0.050
0.050
0.050
0.050
0.050
0.050
0.10
0.10
0.10
0.10
0.10
0.10
0.10
0.50
0.10
0.10
0.050
0.050
5.0
1.0
2.0
1.0
1.0
1.0
1.0
1.0
Soil
UQ/Kq
1.7
1.7
1.7
1.7
1.7
1.7
1.7 -
1.7
3.3
3.3
3.3
3.3
3.3
3.3
3.3
17
3.3
3.3
1.7
1.7
170
33
67
33
33
33
33
33
Limits
On Column
(oa)
5
5
5
5
5
5
5
5
.10
10
10
10
10
10
10
50
10
10
5
5
500
100
200
100
100
100
100
100

       2There is no differentiation between the preparation of low and medium
 soil samples in this method for the analysis of pesticides/Aroclors.

       3Only the exo-epoxy isomer  (isomer B) of heptachlor epoxide is reported
 on the data reporting forms  (Exhibit B).

                                       C-8              '                  OLM04.2

-------
                 Exhibit D - Analytical Methods for Volatiles

                              Table  of  Contents

Section                                                •                   Page

1.0   SCOPE AND APPLICATION 	   4

2.0   SUMMARY OF METHOD	5
      2.1   Water	5
      2.2   Low Level Soil	5
      2.3   Medium Level Soil	5

3.0   DEFINITIONS	5

4 . 0   INTERFERENCES	6

5.0   SAFETY	6

6.0   EQUIPMENT AND SUPPLIES  	   7

7.0   REAGENTS AND STANDARDS	13
      7.1   Reagents	13
      7.2   Standards	13
      7.3   Storage of Standard Solutions 	  16

8.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE  	  18
      8.1   Sample Collection and Preservation  	  18
      8.2   Procedure for Sample Storage  	  18
      8.3   Contract Required Holding Times	 ...  19

9.0   CALIBRATION AND STANDARDIZATION	  20
      9.1   Instrument Operating Conditions 	 ...  20
      9.2   GC/MS Calibration (Tuning) and Ion Abundance  	  22
      9.3   Initial Calibration	  23
      9.4   Continuing Calibration   	  26

10.0  PROCEDURE	29
      10.1  Sample Preparation	29
      10.2  pH Determination  (Water  Samples)   	  35
      10.3  Percent Moisture Determination  	  35

11.0  DATA ANALYSIS AND CALCULATIONS	37
      11.1  Qualitative Identification	37
      11.2  Calculations	39
      11.3  Technical Acceptance Criteria for Sample Analysis 	  44
      11.4  Corrective Action for Sample Analysis  	  45

12.0  QUALITY CONTROL	48
      12.1  Blank Analyses	48
      12.2  Matrix Spike/Matrix Spike Duplicate (MS/MSD)  	  51

13.0  METHOD PERFORMANCE  	  55

14.0  POLLUTION  PREVENTION	55

                                    D-2/VOA                             OLM04.2

-------
15.0  WASTE MANAGEMENT  	 55

16.0  REFERENCES  .	55

17.0  TABLES/DIAGRAMS/FLOWCHARTS	56

APPENDIX A - SCREENING OF HEXADECANE EXTRACTS FOR VOLATILES 	 65
      1.0   SCOPE AND APPLICATION	65
      2.0   SUMMARY OF METHODS	65
      3.0   INTERFERENCES	66
      4.0   SAFETY	66
      5.0   EQUIPMENT AND SUPPLIES	66
      6.0   REAGENTS AND STANDARDS	67
      7.0   QUALITY CONTROL	68
      8.0   CALIBRATION AND  STANDARDIZATION	69
      9.0   PROCEDURE	69

APPENDIX B - MODIFIED SW-846 METHOD 5035 FOR VOLATILES IN LOW LEVEL SOILS  . 72
      1.0   SCOPE AND APPLICATION	'.  .   . 72
      2. 0   SUMMARY OF METHOD	72
      3.0   INTERFERENCES	72
      4.0   SAFETY	,   . 73
      5.0   EQUIPMENT AND SUPPLIES	73
      6. 0   REAGENTS AND STANDARDS	'	75
      7.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE   	 76
      8.0   CALIBRATION AND  STANDARDIZATION	77
      9.0   PROCEDURE	:	79
      10.0  DATA ANALYSIS AND  CALCULATIONS	, .   . 81
      11.0. QUALITY CONTROL	81
                                    D-3/VOA                             OLM04.2

-------
Exhibit D Volatiles -- Section 1
Scope and Application

1.0  SCOPE AND APPLICATION

1.1  In 1978, EPA Headquarters and Regional representatives designed
     analytical methods for the analysis of volatiles in hazardous waste
     samples.  These methods were based on EPA Method 624, Purgeables.  In
     1980, these methods were adopted for use in the Contract Laboratory
     Program  (CLP).  As the requirements of CERCLA evolved, the CLP methods,
     as well as their precedent EPA 600 Series methods, established the basis
     for other EPA methods to perform the analysis of volatiles contained in
     hazardous waste samples  (i.e., Stf-846) .  The following CLP method has
     continuously improved to incorporate technological advancements
     promulgated by EPA, and has continued to set the standard for the
     preparation, extraction, isolation, identification, and reporting of
     volatiles at hazardous waste sites.

1.2  The analytical method that follows is designed to analyze water,
     sediment, and soil from hazardous waste sites for the volatile organic
     compounds on the Target Compounds List (TCL, see Exhibit C).   The method
     includes sample preparation, screening to determine the approximate
     concentration of organic constituents in the sample, and the actual
     analysis which is based on a purge and trap gas chromatograph/mass
     spectrometer  (GC/MS) method.

1.3  This analytical method includes the use of  the Modified SW-846 Method
     5035 for the preparation and analysis of low level soil/sediment samples.
     A detailed description of the sample preparation, analysis, and quality
     control procedures to be followed when this method option is requested
     can be found in Appendix B.

1.4  Problems have been associated with the following compounds analyzed by
     this method.

     •  Chloromethane, vinyl chloride, bromomethane, and chloroethane can
        display peak broadening if the compounds are not delivered to the GC
        column in a tight band.

     •  Acetone, hexanone, 2-butanone, and 4-methyl-2-pentanone have poor
        purge efficiencies.

     •  1,1,1-trichloroethane and all the dichloroethanes can
        dehydrohalogenate during storage or analysis.

     •  Chloromethane can be lost if the purge flow is too fast.

     •  Bromoform is one of the compounds most likely to be adversely affected
        by cold spots and/or active sites in the transfer lines.  Response of
        •its quantitation ion  (m/z 173) is directly affected by the tuning of
        the GC/MS to meet the instrument performance criteria for 4-
        bromofluorobenzene  (BFB) at ions m/z 174/176.  Increasing the m/z
        174/176 ratio may improve bromoform response.
                                    D-4/VOA                            OLM04.2

-------
                                          Exhibit D Volatile -- Sections 2 & 3
                                                 Summary of Method/Definitions

2.0  SUMMARY OF METHOD

2.1  Water

     An inert gas is bubbled through a.5 mL sample contained in a specifically
     designed purging chamber at ambient temperature.   The purgeable compounds
     are efficiently transferred from the aqueous phase to the vapor phase.
     The vapor is swept through-a sorbent column where the purgeables are
     trapped.  After purging is completed, the sorbent column is heated and
     backflushed with the inert gas to desorb the purgeable compounds onto a
     gas chromatographic column.  The gas chromatograph is temperature-
     programmed to separate the purgeable compounds which are then detected
     with a mass spectrometer.

2.2  Low Level Soil

     An inert gas is bubbled through a mixture of reagent water and 5 g of
     sample contained in a specifically designed purging chamber that is held
     at an elevated temperature.  The purgeable compounds are efficiently
     transferred from the aqueous phase to the vapor phase.  The vapor is
     swept through a sorbent column where the purgeables are trapped.  After
     purging is completed, the  sorbent column is heated and backflushed with
     the inert gas to desorb the purgeable compounds onto a gas
     chromatographic column.  The gas chromatograph is temperature-programmed
     to separate the purgeable  compounds which are then detected with a mass
     spectrometer.

2.3  Medium Level Soil

     A measured amount  of  soil  is  collected/extracted with methanol.  A
     portion of the  methanol  is diluted to  5  mL  with  reagent  water.  An inert
     gas  is  bubbled  through this solution in  a specifically designed purging
     chamber at ambient temperature.  The purgeable compounds are effectively
     transferred from the  aqueous  phase to  the vapor  phase.   The vapor  is
     swept  through  a sorbent  column where the purgeables  are  trapped.   After
     purging is completed,  the  sorbent  column is heated and backflushed with
     the  inert gas  to desorb  the purgeable  compounds  onto a gas
     chromatographic column.   The  gas chromatograph is temperature-programmed
     to separate the purgeable  compounds which are then detected with  a mass
     spectrometer.

 3 . 0  DEFINITIONS

     See Exhibit G for  a complete  list  of definitions.
                                    D-5/VOA
OLM04.2

-------
Exhibit D Volatiles -- Sections 4 & b
Interferences/Safety

4.0  INTERFERENCES

4.1  Method interference may be caused by impurities in the purge gas, organic
     compounds out-gassing from the plumbing ahead of the trap, and solvent
     vapors in the laboratory.  The analytical system must be demonstrated to
     be free from contamination under the conditions of the analysis by
     running laboratory method and instrument blanks as described in Section
     12.  The use of non-Polytetrafluoroethylene  (PTFE) tubing, non-PTFE
     thread sealants, or flow controllers with rubber components in the
     purging device should be avoided.

4.2  Samples can be contaminated by diffusion of volatile organics
     (particularly fluorocarbons and methylene chloride) through the septum
     seal into the sample during storage and handling.

4.3  Contamination by carryover can occur whenever high level and low level
     samples are sequentially analyzed.  To reduce carryover, the purging
    'device and sampling syringe must be rinsed with reagent water between
     sample analyses.  For samples containing large amounts of water-soluble
     materials, suspended solids, high-boiling compounds, or high purgeable
     levels, it may be necessary to wash out the purging device with a
     detergent solution between analyses, rinse it with distilled water, and
    .then dry it in an oven at 105 °C.  The trap and other parts of the system
     are also subjected to contamination; therefore, frequent bakeout and
     purging of the entire system may be required.

4.4  The laboratory where volatile analysis is performed should be completely
     free of solvents.                                                    •  '

5.0  SAFETY

5.1  The toxicity or carcinogenicity of each reagent used in this method has
     not been precisely defined; however, each chemical compound should be
     treated as a potential health hazard.  From this viewpoint, exposure to
     these chemicals must be reduced to the lowest possible level by whatever
     means available.  The laboratory is responsible for maintaining a current
     awareness file of OSHA regulations regarding the safe handling of the
     chemicals specified in this method.  A reference file of material data
     handling sheets should also be made available to all personnel involved
     in the chemical analysis.

5.2  The following analytes covered by this method have been tentatively
     classified as known or suspected, human or mammalian carcinogens:
     benzene, carbon tetrachloride, chloroform, and vinyl chloride.  Primary
     standards of these toxic compounds should be prepared in a hood.  A
     NIOSH/Mass approved toxic gas respirator should be worn when the analyst
     handles high concentrations of these toxic compounds.
                                    D-6/VOA                            OLM04.2

-------
                                              Exhibit D Volatiles -- Section 6
                                                        Equipment and Supplies
6.0  EQUIPMENT AND SUPPLIES
     Brand names, suppliers, and part numbers are for illustrative purposes
     only..  No endorsement is implied.  Equivalent performance may be achieved
     using equipment and supplies other than those specified here, but
     demonstration of equivalent performance meeting the requirements of this
     SOW is the responsibility of the Contractor.  The Contractor shall
     document any use of alternate equipment or supplies in the SDG Narrative.

6.1  Glassware

6.1.1   Syringes - 5 mL, gas-tight with shut-off valve.  Micro syringes - 25
        uL and larger, 0.006 inch ID needle.

6.1.2   Syringe Valve - two-way, with Luer ends  (three each), if applicable to
        the purging device.

6.1.3   Pasteur Pipets - disposable.

6.1.4   Vials and Caps - 2 mL for GC.

6.1.5   Volumetric Flasks.

6.1.6   Bottle - 15 mL, screw-cap, with Teflon cap liner.                '

6.2  pH Paper - wide  range

6.3  Balances - analytical,  capable of accurately weighing ± 0.0001 g, and a
     top-loading balance capable of weighing 100 g ± 0.01 g.  The balances
     must be calibrated with class S weights or known reference weights once
     per each 12-hour work  shift.  The balances must be calibrated with class
     S weights at a minimum of  once per month.  The balances must also be
     annually checked by a  certified  technician.

6.4  Purge and Trap Device  - consists of three separate pieces of equipment:
     the sample purge chamber,  trap,  and the desorber.  Several complete
     devices are now  commercially available.

6.4.1   The sample purge chamber must be designed to accept  5 mL  samples with
        a water  column  at  least 3 cm  deep.  The gaseous head space between the
        water column  and the trap must have a total volume of less than 15 mL.
        The purge gas must  pass through the water column as  finely divided
        bubbles, each with a diameter of less than 3 mm at the origin.  The
        purge gas must  be  introduced  no more than 5 mm from  the base of the
        water column.

6.4.2   The trap must be at least 25  cm long and have an inside diameter of at
        least 0.105  inch.   The  trap must be packed to contain the following
        minimum  lengths of absorbents:  (starting from inlet) 0.5  cm silanized
        glass wool,  1 cm methyl silicone,  8 cm of 2,6-diphenylene oxide
        polymer  (Tenax-GC,  60/80 mesh), 8  cm of  silica gel  (Davison Chemical,
        35/60 mesh,  grade  15 or equivalent),  7 cm of coconut charcoal  (prepare
        from Barnebey Cheney, CA-580-26, or equivalent, by crushing  through 26
        mesh screen), and  0.5 cm silanized glass wool.  A description of the
        trap used  for analysis  shall  be provided in the SDG  Narrative.

                                    D-7/VOA                            OLM04.2

-------
Exhibit D Volatiles -- Section 6
Equipment and Supplies

6.4.3   The desorber should be capable of rapidly heating the trap to 180 °C.
        The polymer section of the trap should not be heated higher than
        180 °C and the remaining sections should not exceed 220 °C during
        bakeout mode.

6.4.4   Trap Packing

6.4.4.1   2,6-Diphenylene oxide polymer, 60/80 mesh chromatographic grade
          (Tenax GC or equivalent).

6.4.4.2   Methyl silicone packing, 3.0 percent OV-1 on Chromasorb W, 60/80
          mesh  (or equivalent).

6.4.4.3   Silica gel,  35/60 mesh, Davison, grade 15  (or equivalent)-.

6.4.4.4   Coconut charcoal  (prepare from Barnebey Cheney, CA-580-26, or
          equivalent,  by crushing  through 26 mesh screen).

6.4.4.5   Alternate sorbent traps may be used if:

          •   The trap packing materials do not introduce contaminants which
              interfere with identification and quantitation of the'compounds
              listed in Exhibit C  (Volatiles).

          •   The analytical results generated using the trap meet the initial
              and continuing calibration technical acceptance criteria listed
              in the SOW,and the CRQLs listed in Exhibit C  (Volatiles).

          •   The trap can accept up to 1000 ng of each compound listed in
              Exhibit C  (Volatiles) without becoming, overloaded.

6.4.4.5.1         The alternate trap must be designed to optimize performance.
                  Follow manufacturer's instructions for the use of its
                  product.  Before use of any trap, other than the one
                  specified in Section 6.4.2, the Contractor must first meet
                  the criteria listed in Section 6.4.4.5.  Once this has been
                  demonstrated, the Contractor must document its use in each
                  SDG Narrative by specifying .the trap composition  (packing
                  material/brand name, amount of packing material).  Other
                  sorbent traps include, but are not limited to, Tenax/Silica
                  Gel/Carbon Trap  from EPA Method 524.2, Tenax - GC/Graphpac-D
                  Trap  (Alltech) or~equivalent, and Vocarb 4000 Trap  (Supelco)
                  or equivalent.

6.4.4.5.2         The Contractor must maintain documentation that the
                  alternate trap meets the criteria listed in Section 6.4.4.5.
                  The minimum documentation requirements are as follows:

6.4.4.5.2.1          Manufacturer provided information concerning the
                     performance characteristics of the trap.

6.4.4.5.2.2          Reconstructed ion chromatograms and data system reports
                     generated on  the Contractor's GC/MS used for CLP
                     analyses:
                                    D-8/VOA                            OLM04.2

-------
                                                Exhibit D Volatiles -- Section 6
                                                          Equipment and Supplies

                       •  From instrument blank analyses which demonstrate that
                          there are no contaminants which interfere with the
                          volatile analysis when using the alternate trap;

                       •  From initial and continuing calibration standards
                          analyzed using the trap specified in Section 6.4.4.

|  6.4.4.5.2.3          Based on Contractor generated data described above, the
                       Contractor must complete a written comparison/review,
                       which has been signed by the Laboratory Manager,
                       certifying that:

                       •  The alternate trap performance meets the technical
                          acceptance criteria listed in Sections 9.3.5 and
                          9.4.5;

                       •  The low point initial calibration standard analysis
                          has adequate sensitivity to meet the volatile CRQLs;

                       •  The high point initial calibration standard analysis
                          was not overloaded;

                       •  The alternate trap materials do not introduce
                          contaminants which interfere with the identification
                          and/or quantitation of the compounds listed in Exhibit
                          C  (Volatiles).

|  6.4.4.5.2.4          The documentation must be made available to the Agency
                       during on-site  laboratory evaluations or sent to the
                       Agency upon request of the Technical Project Officer  or
                       the Administrative Project Officer.

  6.4.5   The purge and trap apparatus may be assembled as a separate unit or be
          an integral unit coupled with a gas chromatograph.

  6.5  A heater or heated bath  capable of maintaining the purge chamber at 40 °C
        (± 1 °C) is to be used for low level soil/sediment analysis, but not  for
       water or medium level soil/sediment analyses.

  6.6  Gas Chromatography/Mass  Spectrometer  (GC/MS) System

  6.6.1   Gas Chromatograph  - the gas  chromatograph  (GC) system must be capable
          of temperature programming and have a flow controller that maintains a
          constant column flow  rate throughout desorption and temperature
          program operations.   The system must include or be interfaced to a
          purge and trap system as specified in Section  6.4 and have all
          required accessories  including syringes, analytical columns, and
          gases.  All GC carrier gas lines must be constructed from stainless
          steel or copper tubing.  Non-PTFE thread sealants, or flow controllers
          with rubber components are not to be used.

  6.6.2   Gas Chromatography Columns  -  a description of  the column used for
          analysis shall be  provided in the SDG Narrative.

  6.6.2.1    Packed columns  - 6  ft. long x  0.1  inch ID glass, packed with  1.0

                                     ..D-9/VOA                             OLM04.2

-------
Exhibit D Voiatiles -- Section 6
Equipment and Supplies

          percent SP-1000 on Carbopack B (60/80) mesh or equivalent.

6.6.2.2   Capillary Columns

6.6.2.2.1     Minimum length 30 m x 0.53 mm ID VOCOL  (Supelco) or equivalent
             . fused silica widebore capillary column with 3 um film thickness.

6.6.2.2.2     Minimum length 30 m x 0.53 mm ID DB-624  (J & W Scientific) or
              equivalent fused silica widebore capillary column with 3 pm film
              thickness.'

6.6.2.2.3     Minimum length 30 m x 0.53 mm ID AT-624  (Alltech) or equivalent
              fused silica widebore capillary column with 3 um film thickness.

6.6.2.2.4     Minimum length 30 m x 0.53 mm ID HP-624  (Hewlett-Packard) or
              equivalent fused silica widebore capillary column with 3 um film
              thickness.

6.6.2.2.5     Minimum length 30 m x 0.53 mm ID RTx-624  (Restek) or equivalent
              fused silica widebore capillary column with 3 um film thickness.

6.6.2.2.6     Minimum length 30 m x 0.53 mm ID BPX-624  (SGE) or equivalent
              fused silica widebore capillary column with 3 um film thickness.
                               ->
6.6.2.2.7     Minimum length 30 m x 0.53 mm ID CP-Sil  13CB  (Chrompack) or
              equivalent fused silica widebore capillary column with 3 pm film
              thickness.

6.6.2.3   A capillary column is considered equivalent  if:

          •   The column does not introduce contaminants which interfere with
              the identification and quantitation of the compounds listed in
              Exhibit C  (Voiatiles).

          •   The analytical results generated using the column meet the
              initial and continuing calibration technical acceptance criteria
              listed in the SOW, and the CRQLs listed  in Exhibit C
               (Voiatiles).

          •   The column can accept up to 1000 ng of each compound listed in
              Exhibit C  (Voiatiles) without becoming overloaded.

          •   The column provides equal or better resolution of the compounds
              listed in Exhibit C  (Voiatiles) than the  columns listed in
              Section 6.6.2.2.

6.6.2.4   As applicable, follow the manufacturer's instructions for use of its
          product.

6.6.2.5   The Contractor must maintain documentation that the column met the
          criteria in Section 6.6.2.3.  The minimum documentation is as
          follows:

6.6.2.5.1     Manufacturer provided information concerning the performance
              characteristics of the column.

                                   D-10/VOA                            OLM04.2

-------
                                              Exhibit D Volatiles -- Section 6
                                                        Equipment and Supplies

6.6.2.5.2     Reconstructed ion chromatograms and data system reports
              generated on the GC/MS used for the CLP analyses:

              •   From instrument blanks which demonstrate that there are no
                  contaminants which interfere with the volatile analysis when
                  using the alternate column,-

              •   From initial and continuing calibration standards analyzed
                  using the alternate column.

6.6.2.5.3     Based on the Contractor-generated data described above, the
             ' Contractor shall complete a written review, signed by the
              Laboratory Manager, certifying that:

              •   The column performance meets the technical acceptance
                  criteria in Sections 9.3:5 and 9.4.5;

              •   The low point initial calibration standard analysis has
                  adequate sensitivity to meet the volatile CRQLs;

              •   The high point initial calibration standard analysis was not
                  overloaded;

              •   The column does not introduce contaminants which interfere
                  with the identification and/or quantitation of compounds
                  listed in Exhibit C  (Volatiles).

6.6.2.5.4     The documentation must be made available to the Agency during
              on-site laboratory evaluations or sent to  the Agency upon
              request of the Technical Project Officer or Administrative
              Project Officer.

6.6.3   Mass Spectrometer  - must be capable of scanning  from 35 to 300 amu
        every 1  second or  less to every 2 seconds or less utilizing  70 volts
         (nominal) electron energy in the electron impact ionization mode, and
        producing a mass spectrum which meets all the instrument performance
        acceptance criteria when 50 ng of BFB is injected through the gas
        chromatograph inlet.  The instrument conditions  required for the
        acquisition of the BFB mass spectrum are given in Section 9.

6.6.3.1   NOTE:  The MS scan rate should allow acquisition of at least five
          spectra while a  sample compound elutes from the GC.  The purge and
          trap GC/MS system must be in a room whose atmosphere is demonstrated
          to be  free of all potential contaminants which will interfere with
          the analysis.  The instrument must be vented to the outside of the
          facility or to a trapping system which prevents the release of
          contaminants into the  instrument room.

6.6.4   GC/MS interface  -  any gas chromatograph to mass  spectrometer interface
         that gives acceptable calibration points at 50 ng or less per
         injection for each of the parameters of interest, and achieves all
         acceptance criteria, may be used.  Gas chromatograph to mass
         spectrometer interfaces  constructed  of all-glass or glass-lined
        materials are recommended.  Glass can be deactivated by  silanizing
         with dichlorodimethylsilane.

                                    D-ll/VOA                            OLM04.2

-------
Exhibit D Volatilss -- Section 6
Equipment and Supplies

6.6.5   Data system - a computer system interfaced to the mass spectrometer
        that allows the continuous acquisition and storage, on machine-
        readable media, of all mass spectra obtained throughout the duration
        of the chromatographic program.  The computer must have software that
        allows searching any GC/MS data file for ions of a specified mass and
        plotting such ion abundances versus time or scan number.  This type of
        plot is defined as an Extracted Ion Current Profile (EICP).   Software
        must also be available that allows integrating the abundance in any
        EICP between specified time or scan number limits.  Also, for the non-
        target compounds, software must be available that allows for the
        comparison of sample spectra against reference library spectra.  The
        NIST/EPA/NIH  (May 1992 release or later) and/or Wiley (1991 release or
        later), or equivalent mass spectral library shall be used as the
        reference library.  The data system must be capable of flagging all
        data files that have been edited manually by laboratory personnel.

6.6.6   Magnetic tape storage device - capable of recording data and must be
        suitable for long-term, off-line storage.
                                   D-12/VOA                            OLM04.2

-------
                                              Exhibit D Volatiles -- Section 7
                                                        Reagents and Standards

7.0  REAGENTS AND STANDARDS

7.1  Reagents

7.1.1   Reagent water - defined as water in which an interferant is not
        observed at or above the CRQL of the analytes of interest.  Reagent
        water may be generated by passing tap water through a carbon filter
        bed containing about 453 g  (1 Ib) of activated carbon (Calgon Corp.,
        Filtrasorb-300 or equivalent).

7.1.1.1   A water purification system (Millipore Super-Q or equivalent) may be
          used to generate reagent water.

7.1.1.2   Reagent water may also be prepared by boiling water for 15 minutes.
          Subsequently, while maintaining the temperature at 90 °C,  bubble a
          contaminant-free inert gas through the water for one hour.  While
          still hot, transfer the water to a narrow-mouth screw-cap bottle and
          seal with a Teflon-lined septum and cap.

7.1.2   Methanol - pesticide quality or equivalent

7.2  Standards
7.2.1   Introduction

        The Contractor must provide all standards to be used with this
        contract.  These  standards may be used only after they have been
        certified according to the procedure in Exhibit E.  The Contractor
        must be  able to verify that the standards are certified.
        Manufacturer's certificates of analysis must b.e retained by the
        Contractor and presented upon request.

7.2.2   Stock  Standard Solutions

7.2.2.1   Stock  standard  solutions may be purchased or may be prepared in
          tnethanol from pure  standard materials.

7.2.2.2   Prepare stock standard solutions by placing about 9.8 mL of methanol
          into a 10 mL ground-glass stoppered volumetric flask.  Allow the
          flask  to stand,  unstoppered, for about 10 minutes, or until all .
          alcohol-wetted  surfaces have dried.  Weigh the flask to the nearest
          0.1  mg.

7.2.2.3   Add  the assayed reference material as described below.  .

7.2.2.3.1      If the compound is a liquid, using a 100 uL syringe, immediately
               add two or  more drops of assayed reference material to the
               flask, then reweigh.  The liquid must fall directly into the
               alcohol without contacting  the neck of the flask.

7.2.2.3.2      If the compound is a gas at room temperature, fill a 5 mL valved
               gas-tight syringe with the  reference standard to the 5 mL mark.
               Lower the needle to 5 mm above the methanol meniscus.  Slowly
               introduce the  reference standard above the surface of the
               liquid.  The gas will rapidly dissolve in the methanol.
                                    D-13/VOA
OLM04.2

-------
Exhibit D Volatiles -- Section 7
Reagents and Standards

7.2.2.3.3     The procedure in Section 7.2.2.3.2 may also be accomplished by
           .  . using a lecture bottle equipped with a Hamilton Lecture Bottle
              Septum  (#86600).  Attach Teflon tubing to the side-arm relief
              valve and direct a gentle"stream"of the reference standard"into
              the methanol meniscus.

7.2.2.3.4     Reweigh, dilute to volume,  stopper, then mix by inverting the
              flask several times.  For non-gaseous compounds, calculate the
              concentration in micrograms per microliter from the net gain in
              weight.   When compound purity is assayed to be 97.0 percent or
              greater, the weight may be used without correction to calculate
              the concentration of the stock standard.  If the compound purity
              is assayed to be less than 97.0 percent, the weight must be
              corrected when calculating the concentration of the stock
              solution.  See Exhibit E (Analytical Standards Requirements).
              For gaseous -compounds, calculate the concentration in micrograms
              per microliter, using the Ideal Gas Law; taking into account the
              temperature and pressure conditions within the laboratory.

7.2.2.3.5     Prepare fresh stock standards every two months for gases or for
              reactive compounds such as styrene.  All other stock standards
              for non-gases/non-reactive purgeable compounds must be replaced
              after six months, or sooner if the standard has degraded or
              evaporated.

7.2.3-  Secondary Dilution Standards

7.2.3.1   Using stock standard solutions, prepare secondary dilution standards
      "    in methanol that contain the compounds of interest, either singly or
      -    mixed together.  Secondary dilution standard solutions should be
          prepared at concentrations that can be easily diluted to prepare
          working standard solutions.

7.2.3.2   Prepare fresh secondary dilution standards for gases and for
          reactive compounds such as styrene every month, or sooner, if
          standard has degraded or evaporated.  Secondary dilution standards
          for the other purgeable compounds must be replaced after six months,
          or sooner, if standard has degraded or evaporated.

7.2.4   Working Standards

7.2.4.1   System Monitoring Compound  (SMC) Spiking Solution

          Prepare a system monitoring compound spiking solution containing
          toluene-d8,  4-bromofluorobenzene (BFB), and 1,2-dichloroethane-d4 in
          methanol at a concentration of 25 ug/mL.  Add 10 uL of this spiking
          solution into 5 mL of sample, sample extract or calibration standard
          for -a concentration of 50 ug/L.  Prepare fresh spiking solution
          weekly, or sooner, if the solution has degraded or evaporated.

7.2.4.2   Matrix Spiking Solution

          Prepare a spiking solution in methanol•that contains the following
          compounds at a concentration of 25 ug/mL:  1,1-dichloroethene,
          trichloroethene, chlorobenzene, toluene, and benzene.  Prepare fresh

                                   D-14/VOA                            OLM04.2

-------
                                              Exhibit D Volatiles -- Section 7
                                                        Reagents and Standards

          spiking solution weekly,  or sooner,  if the solution has degraded or
          evaporated.

7.2.4.3   Internal Standard Spiking Solution

          Prepare an internal standard spiking solution containing
          bromochloromethane, chlorobenzene-d5, and 1,4-difluorobenzene in
          methanol at a concentration of 25 ug/mL for each internal standard.
          Add 10 uL of this spiking solution into 5 mL of sample or
          calibration standard for a concentration of 50 ug/L.  Prepare fresh
          spiking solution weekly,  or sooner,  if the solution has degraded or •
          evaporated.

7.2.4.4   Instrument Performance Check Solution - 4-Bromofluorobenzene (BFB)

          Prepare a 25 ng/uL solution of BFB in methanol.  Prepare fresh BFB
          solution every six months, or sooner, if the solution has degraded
          or evaporated.  NOTE:  The 25 ng/uL concentration is used with a 2
          uL injection volume.  The laboratory may prepare a 50 ng/uL solution
          of BFB if a 1 uL injection volume" is used.

7.2.4.5   Calibration Standard.Solution

          Prepare a calibration standard solution containing all of the
          purgeable target compounds in methanol.  The recommended
          concentration of the target compounds is 100 ug/mL.  Prepare fresh
          calibration standard solutions weekly, or sooner, if solutions have
          degraded or evaporated.

7.2.4.6   Initial and Continuing Calibration Standards

7.2.4.6.1     Prepare five aqueous  initial  calibration standard solutions
              containing all of the purgeable  target  compounds and system
              monitoring compounds  at 10, 20,  50, 100, and 200 ug/L levels.
              It is  required that all three xylene isomers  (o-, m-, and p-
              xylene) be present in the calibration standards at
              concentrations of each isomer equal to  that of the other target
              compounds.(i.e., 10,  20, 50,  100, and 200 ug/L).  Similarly,  the
              cis and trans  isomers of 1,2-dichloroethene must both be present
              in the standards at concentrations of each isomer equal to that
              of the other target compounds.

7.2.4.6.2     Aqueous calibration standards may be prepared in a volumetric
              flask  or  in the  syringe used  to  inject  the standard  into the
              purging device.

7.2.4.6.2.1       Volumetric flask  - add an appropriate volume of  the 100
                  ug/mL calibration standard  solution (Section 7-2.4.5) to  an
                  aliquot of reagent water  in  a volumetric flask.  Use a
                  microsyringe and  rapidly  inject the alcohol standard into
                  the expanded area of  the  filled volumetric  flask.  Remove
                  the needle as quickly as  possible  after  injection.  Bring.to
                  volume.  Mix by  inverting the  flask three times  only.
                  Discard the  contents  contained in  the head  of  the flask.
                                    D-15/VOA                            OLM04.2

-------
Exhibit D Volatiles -- Section 7
Reagents and Standards

7.2.4.6.2.2       Syringe - remove the plunger from a 5 mL "Luerlock" syringe.
                  Pour reagent water into the syringe barrel to just short of
                  overflowing.  Replace the syringe plunger and compress the
                  water.  Invert the syringe, open the syringe valve and vent
                  any residual air.  Adjust the water volume to 5 mL minus the
                  amount of calibration standard to be added.  Withdraw the
                  plunger slightly and add an appropriate volume of working
                  calibration standard through the valve bore of the syringe.
                  Close the valve and invert three times.

7.2.4.6.2.3       The 50 ug/L aqueous calibration standard solution is the
                  continuing calibration standard.

7.2.4.6.3     The methanol contained in each of the aqueous calibration
              standards must not exceed 1.0 percent by volume.

7.2.5   Ampulated Standard Extracts

        Standard solutions purchased from a chemical supply house as ampulated
        extracts in glass vials may be retained and used until the expiration
        date provided by the manufacturer.  If no manufacturer's expiration
        .date is provided, the standard solutions as ampulated extracts may be
        retained and used for 2 years from the preparation date.  Standard
      -r-  solutions, prepared by the Contractor which are immediately ampulated
      -  in glass vials, may be retained for 2 years from the preparation date.
      /;  Upon breaking the glass seal, the expiration times listed in Sections
      •••  7.2.2 through 7.3 will apply.  The Contractor is responsible for
      '  assuring that the integrity of the standards has not degraded  (see
        Section 7.3.5).

7.3  Storage of Standard Solutions

7.3.1   Store the stock standards in Teflon-sealed screw-cap bottles with zero
        headspace at -10 °C to -20 °C,  and protect the standards from light.
        Once one of the bottles containing the stock standard solution has
        been opened, it may be used for no longer than one week.

7.3.2   Store secondary dilution standards in Teflon-sealed screw-cap bottles
        with minimal headspace at -10 °C to -20 °C,  and protect the standards
        from light.  The secondary dilution standards must be checked
        frequently for signs of degradation or evaporation, especially just
        prior to preparing working standards from them.

7.3.3   Aqueous standards may be stored for up to 24 hours if held in
        Teflon-sealed screw-cap vials with zero headspace at 4 °C (±2 °C) .
        Protect the standards from light.  If not so stored, they must be
        discarded after one hour unless they are set up to be purged by an
        autosampler.  When using an autosampler the standards may be kept for
        up to 12 hours in purge tubes connected via the autosampler to the
        purge and trap device.  All other working standards may be stored at
        -10 °C to -20 °C.

7.3.4   Purgeable standards must be stored separately from other standards.
                                    D-16/VOA                            OLM04.2

-------
                                              Exhibit D Volatiles -- Section 7
                                                        Reagents and Standards

7.3.5   The Contractor is responsible for maintaining the integrity of
        standard solutions and verifying prior to use.  This means that
        standards must be brought to room temperature prior to use,  checked
        for losses, and checked that all components have remained in the
        solution.
                                    D-17/VOA                           OLM04.2

-------
Exhibit D Volatiles -- Section 8
Sample Collection, Preservation, and Storage

8.0  SAMPLE COLLECTION, PRESERVATION, AND STORAGE

8.1  Sample Collection and Preservation

8.1.1   Water samples may be collected in glass containers having a total
        volume of at least 40 mL with a Teflon-lined septum and an open top
        screw-cap.  Soil samples may be collected in glass containers or
        closed end tubes (e.g., brass sleeves) in sufficient quantity to
        perform the analysis.  Headspace should be avoided.  Soil samples for
        medium level analysis may also be collected in pre-weighed vials
        containing 10 ml of methanol.  The specific requirements for site
        sample collection are outlined by the Region.

8.1.2   For collection of water samples, the containers must be. filled in such
        a manner that no air bubbles pass through the sample as the container
        is being filled.  Seal the vial so that no air bubbles are entrapped
        in it.

8.1.3   Water samples are preserved to a pH of 2 at the time of collection.

8.1.4   For collection of medium level soil samples with methanol, the sample
        vial, with 10 ml of methanol and all labeling, is weighed to the
        nearest 0.1 g prior to the addition of sample.  Approximately 5 g of
        sample is added to the vial.  The sample vial with sample is weighed
        to the nearest 0.1 g.  The initial weight, final weight and sample
        weight will be recorded and provided to the laboratory.

8.1.5   All samples must be iced or refrigerated at 4 °C (±2 °C)  from the time
        of collection until analysis.

8.2  Procedure for Sample Storage

8.2.1   The samples must be protected from light and refrigerated at 4 °C
        (±2 °C)  from the time of receipt until 60 days after delivery of a
        reconciled, complete sample data package to the Agency.  After 60
        days, the samples may be disposed of in a manner that complies with
        all applicable regulations.

8.2.2   The samples must be stored in an atmosphere demonstrated to be free of
        all potential contaminants and in a refrigerator used only for storage
        of volatile samples.

8.2.3   All volatile samples in an SDG'must be stored together in the same
        refrigerator.

8.2.4   Storage blanks shall be stored with samples until all samples are
        analyzed.

8.2.5   Samples, sample extracts, and standards must be stored separately.

8.2.6   Volatile  standards must be  stored separately  from  semivolatile and
        pesticide/Aroclor standards.
                                   D-18/VOA                            OLM04.2

-------
                                              Exhibit D Volatiles -- Section 8
                                  Sample Collection,  Preservation,  and Storage

8.3  Contract Required Holding Times

     Analysis of water and soil/sediment samples must be completed within 10
     days of Validated Time of Sample Receipt (VTSR).   As part of the Agency's
     QA program, the Agency may provide Performance Evaluation (PE)  samples as
     standard extracts which the Contractor is required to prepare per the
     instructions provided by the Agency.  PE samples must be prepared and
     analyzed concurrently with the samples in the SDG.  The contract required
     10 day holding time does not apply to PE samples received as standard
     extracts.
                                    D-19/VOA                           OLM04.2

-------
Exhibit D Volatiles -- Section 9
Calibration and Standardization
Instrument Operating Conditions

9.0  CALIBRATION AND STANDARDIZATION

9.1  Instrument Operating Conditions

9.1.1   Purge and Trap

9.1.1.1   The following are the recommended purge and trap analytical
          conditions.  The conditions are recommended unless otherwise noted.

              Purge conditions

              Purge Gas:                  Helium or Nitrogen
              Purge Time:                 11.0 ± 0.1 minute
              Purge Flow Rate:            25-40 mL/minute
              Purge Temperature:          Ambient temperature for water or
                                          medium level soil/sediment samples
                                           (required); 40 °C low level
                                          soil/sediment samples  (required)

              Desorb Conditions

              Desorb Temperature:         180 °C
              Desorb Flow Rate:           15 mL/minute
              Desorb Time:                4 . 0 ± 0 .1 minute

              Trap Reconditioning Conditions

              Reconditioning Temperature: 180 °C
              Reconditioning Time:        7.0 ± 0.1 minute  (minimum).  A
                                          longer time may be required to bake
                                          contamination or water from the
                                          system.

9.1.1.2   Before initial use, condition the trap overnight at 180 °C by
          backflushing with at least 20 mL/minute flow of inert gas.  Do not
          vent the trap effluent onto the analytical column.  Prior to daily
          use, condition the trap by heating at 180 °C for 10 minutes while
          backflushing.  The trap may be vented to the analytical column
          during daily conditioning; however, the column must be run through
          the temperature program prior to the analysis of samples.

9.1.1.3   Optimize purge and trap conditions for sensitivity and to minimize
          cross-contamination between samples.  Once optimized, the same purge
          and trap conditions must be used for the analysis of all standards,
          samples, and blanks.

9.1.1.4   A moisture reduction/water management system may be used to improve
          the chromatographic performance by controlling moisture or water if:

          •   The system does not introduce contaminants which interfere with
              identification and quantitation of compounds listed in Exhibit C
               (Volatiles),

          •   The analytical results generated when using the moisture

                                   D-20/VOA                            OLM04.2

-------
                                              Exhibit D Volatiles -- Section 9
                                               Calibration and Standardization
                                               Instrument Operating Conditions

              reduction/water management system meet the initial and
              continuing calibration technical acceptance criteria listed in
              the SOW and the CRQLs listed in Exhibit C (Volatiles),

          •   All calibration standards and samples, including blanks and
              MS/MSDs,  are analyzed under the same conditions,

          •   The Contractor performs acceptably on the Performance Evaluation
              samples using this system.

9.1.2   Gas Chromatograph

9.1.2.1   The following are the recommended GC analytical conditions.  These
          conditions are recommended unless otherwise noted.

              Packed columns

              Carrier Gas:                Helium
              Flow Rate:                  30 mL/minute
              Initial Temperature:        45 °C
              Initial Hold Time:          3 minutes
              Ramp Rate:                  8 °C/minute
              Final Temperature:          220 °C
              Final Hold Time:            Until three minutes after all
                                          compounds listed in Exhibit C
                                           (Volatiles) elute  (required)
              Transfer Line Temperature:  250-300 °C

              Capillary Columns

              Carrier Gas:        '        Helium
              Flow Rate:                  15 mL/minute
              Initial Temperature:        10 °C
              Initial Hold Time:          1,0 - 5.0  (± 0.1) minutes
              Ramp Rate:                  6 °C/minute
              Final Temperature:          160 °C
              Final Hold Time:            Until three minutes after all
                                          compounds listed in Exhibit C
                                           (Volatiles) elute  (required)

9.1.2.2   Optimize GC conditions for analyte separation and sensitivity.  Once
          optimized, the same GC conditions must be used for the analysis of
          .all standards, samples, blanks, matrix spikes, and matrix spike
          duplicates.

9.1.2.3   For capillary columns, if the gaseous compounds chloromethane,
          bromomethane, vinyl chloride, and chloroethane fail to exhibit
          narrow, symmetrical peak shape, are not separated  from the solvent
          front, or are not resolved greater than 90.0 percent  from each
          other, then a subambient oven controller must be used, and the
          initial temperature must be  less .than or equal to  10  °C.
                                   D-21/VOA                             OLM04.2

-------
Exhibit D Volatiles -- Section 9
Calibration and Standardization
GC/MS Calibration and Ion Abundance

9.1.3   Mass Spectrometer

        The following are the required mass spectrometer analytical
        conditions:

        Electron Energy:      70 volts  (nominal)
        Mass Range:           35-300 amu
        Scan Time:            To give at least 5 scans per peak, not to exceed
                              2 seconds per scan for capillary column.

                              To give at least 5 scans per peak, not to exceed
                              3 seconds per scan for packed column.

9.2  GC/MS  Calibration  (Tuning) and  Ion Abundance

9.2.1   Summary of GC/MS Performance Check

9.2.1.1   The GC/MS system must be tuned to meet the manufacturer's
          specifications, using a suitable calibrant such as perfluoro-tri-n-
          butylamine  (FC-43) or perfluorokerosene  (PFK).  The mass calibration
          and resolution of the GC/MS system are verified by the analysis of
          the instrument performance check solution  (Section 7.2.4.4).

9.2.1.2   Prior to the analysis of any  samples, including MS/MSDs, blanks, or
          calibration standards, the Contractor must establish that the GC/MS
          system meets the' mass spectral ion abundance criteria for the
          instrument performance check  solution containing BFB.

9.2.2   Frequency of GC/MS Performance  Check

        The instrument performance check solution must be injected once at the
        beginning of each 12-hour period, during which samples or standards
        are to be analyzed.  The twelve  (12) hour time period for GC/MS
        instrument performance check  (BFB), standards calibration  (initial or
        continuing calibration criteria), blank and sample analysis begins at
        the moment of injection of the  BFB analysis that the laboratory
        submits as documentation of a compliant instrument performance check.
        The time period ends after twelve  (12) hours have elapsed according to
        the system clock.

9.2.3   Procedure for GC/MS Performance Check

9.2.3.1   The analysis of the instrument performance check solution may be
          performed as follows:

          •   As an injection of up to  50 ng of BFB into the GC/MS.

          •   By adding 50 ng of BFB to 5 mL of reagent water and analyzing
              the resulting solution as if it were an environmental sample
               (See Section 10).

9.2.3.2   The instrument performance check solution must be analyzed alone
          without calibration standards.  NOTE:  The calibration standards
          contain BFB as a system monitoring compound  (SMC).

                                   D-22/VOA                            OLM04.2

-------
                                              Exhibit D Volatiles -- Section 9 .
                                               Calibration and Standardization
                                                           Initial Calibration

9.2.4   Technical Acceptance Criteria for GC/MS Performance Check

9.2.4.1   The mass spectrum of BFB must be acquired in the following manner.
          Three scans (the peak apex scan and the scans immediately preceding
          and following the apex)  are acquired and averaged.  Background
          subtraction is required, and must be accomplished using a single
          scan no more than 20 scans prior to the elution of BFB.  Do not
          background subtract part of the BFB peak.

9.2.4.2   NOTE:  All subsequent standards, samples, MS/MSD, and blanks
          associated with a BFB analysis must use identical mass spectrometer
          instrument conditions.

9.2.4.3   The analysis of the instrument performance check solution must meet
          the ion abundance criteria given in Table 1.

9.2.5   Corrective Action for GC/MS Performance Check

9.2.5.1   If the technical acceptance criteria are not met, retune the GC/MS
          system.  It may also be necessary to clean the ion source, clean the
          quadrupole rods, or take other corrective actions td achieve the
          technical acceptance criteria.

9.2.5.2   BFB technical acceptance criteria must be met before any standards,
          samples, including MS/MSDs or required blanks are analyzed.  Any
          samples or required blanks analyzed when tuning technical acceptance
          criteria have not been met will require reanalysis at no additional
          cost to the Agency.

9.3   Initial  Calibration

9.3.1   Summary of Initial Calibration

        Prior to the analysis of  samples and required blanks, and after the
        instrument performance check solution criteria have been met, each
        GC/MS system must be calibrated at five concentrations to determine
        instrument sensitivity and the linearity of GC/MS response for the
        purgeable target compounds.

9.3.2   Frequency of Initial Calibration

9.3.2.1   Each GC/MS system must  be calibrated upon award of the contract,
          whenever the Contractor takes corrective action which may change  or
          affect the initial calibration criteria  (e.g., ion source cleaning
          or  repair, column replacement, etc.), or if the  continuing
          calibration acceptance  criteria have not been met.

9.3.2.2   If  time remains  in the  12-hour time period after meeting the
          technical acceptance criteria for the initial calibration, samples
          may be analyzed.  It is not necessary to analyze  a continuing
          calibration standard if the initial  calibration  standard that is  the
          same concentration as the continuing calibration  standard meets the
          continuing calibration  technical acceptance criteria.  A method
          blank  is required.  Quantify all sample  and quality control sample

                                    D-23/VGA                            OLM04.2

-------
Exhibit D Volatiles -- Section 9
Calibration and -Standardization
Initial Calibration

          results, such as internal standard area response change and
          retention time shift, against the initial calibration standard that
          is the same concentration as the continuing calibration standard.

9.3.3   Procedure for Initial Calibration

9.3.3.1   Assemble a purge and trap device that meets the specifications in
          Section 6.4.  Condition the device as described in Section 9.1.1.

9.3.3.2   Connect the purge and trap device to the gas chromatograph.  The
          gas chromatograph must be operated using temperature and flow rate
          parameters equivalent to those in Section 9.1.2.

9.3.3.3   Add 10 uL of the internal standard solution (Section 7.2.4.3) to
          each of the five aqueous calibration standard solutions containing
          the system monitoring compounds  (Section 7.2.4.6) for a
          concentration of 50 pg/L at time of purge.  Analyze each calibration
          standard according to Section 10.

9.3.3.4   Separate initial and continuing calibrations must be performed for
          water samples and low level soil/sediment samples (unheated purge
          vs. heated purge).  Extracts of medium level soil/sediment samples
          may be analyzed using the calibrations of water samples.

          The laboratory may run different matrices in the same 12-hour time
          period under the same tune, as long as separate calibrations are
          performed for each matrix within that 12-hour period.

9.3.4   Calculations for Initial Calibration

9.3.4.1   Calculate the relative response factor  (RRF) for each volatile
          target and system monitoring compound using Equation 1.  The primary
          characteristic ions used for quantitation are listed in Table 2 and
          Table 4.  Assign the target compounds and system monitoring compound
          to an internal standard according to Table 3.  If an interference
          prevents the use of a primary ion for a given internal standard, use
          a secondary ion listed in Table 4.  NOTE:  Unless otherwise stated,
          the area response of the primary characteristic ion is the
          quantitation ion.
          EQ.
                                RRF =    X
                                      Ais    CX
          Where,

          AX   =   Area of the characteristic ion  (EICP) for the compound to be
                  measured  (see Table 2)
          AiS  =   Area of the characteristic ion  (EICP) for the specific
                  internal  standard  (see Tables 3 and 4)
          Cis  =   Concentration of the  internal standard
          Cx   =   Concentration of the  compound to be measured

                                   D-24/VGA                            OLM04.2

-------
                                              Exhibit D Volatiles -- Section 9
                                               Calibration and Standardization
                                                           Initial Calibration
9.3.4.2   Calculating the relative response factor of the xylenes and the bis
          and trans isomers of 1,2-dichloroethene requires special attention.
          On packed columns, o- and p-xylene isomers co-elute.  On capillary
          columns,  the m- and p-xylene isomers co-elute.  Therefore,  when
          calculating the relative response factor in the equation above, use
          the area response (Ax)  and concentration (Cx) of the peak that
          represents the single isomer on the GC column used for analysis.

9.3.4.3   The mean relative response factor (~RR~F~) must be calculated for all
          compounds.

9.3.4.4   Calculate the % Relative Standard Deviation  (%RSD) of the RRF values
          over the working range of the curve.
          EQ.  2

                               Standard Deviation
                        %RSD =
                                      Mean
          Where,
          Standard Deviation =
•£ (x, -x):
i • 1	
   (n-1)
          X±  =   each individual value used to calculate the mean
          X   =   the mean of n values
          n   =   the total number of values

9.3.5   Technical Acceptance Criteria' for Initial Calibration

9.3.5.1   All initial calibration standards must be analyzed at the
          concentration levels described in Section 7.2.4.6.1, and at the
          frequency described in Section 9.3.2 on a GC/MS system meeting the
          BFB technical acceptance criteria.

9.3.5.2   The relative response factor  (RRF) at each calibration concentration
          for each purgeable target and system monitoring compound must be
          greater than or equal to the compound's minimum acceptable response
          factor listed in Table 5.

9.3.5.3   The %RSD for each target or system monitoring compound listed in
          Table 5 must be less than or equal to that value  listed.

9.3.5.4   Up to two compounds may fail the criteria listed  in Sections 9.3.5.2
          and 9.3.5.3 and still meet the minimum response factor and %RSD
          requirements.  However, these compounds must have a minimum RRF
          greater than or equal to 0.010, and the %RSD must be less than or
                                    D-25/VOA                            OLM04.2

-------
Exhibit L) Voiatiies -- section 9
Calibration and Standardization
Continuing Calibration

          equal to 40.0 percent.
9.3.5.5   Excluding those ions in the solvent front,  and the combined xylenes
          in the 200 ug/L standard, no quantitation ion may saturate the
          detector.  Follow the manufacturer's instrument operating manual to
          determine how saturation is indicated for your instrument.

9.3.6   Corrective Action for Initial Calibration

9.3.6.1   If the initial calibration technical acceptance criteria are not
          met, inspect the system for problems.  It may be necessary to clean
          the ion source, change the column, service the purge and trap
          device, or take other corrective actions to achieve the technical
          acceptance criteria.

9.3.6.2   Initial calibration technical acceptance criteria must be met before
          any samples or required blanks are analyzed.  Any samples including
          MS/MSD or required blanks analyzed when initial calibration
          technical acceptance criteria have not been met will require
          reanalysis at no additional cost to the Agency.

9.4  Continuing  Calibration

9.4.1   Summary of Continuing Calibration

        Prior to the analysis of samples and required' blanks and after BFB and
        initial calibration acceptance criteria have been met, each GC/MS
        system must be routinely checked by analyzing a continuing calibration
        standard containing all the purgeable target and system monitoring
        compounds to ensure that the instrument continues to meet the
        instrument sensitivity and linearity requirements of the SOW.

9.4.2   Frequency of Continuing Calibration

9.4.2.1   A check of the calibration curve must be performed once every 12
          hours  (see Section 9.2.2 for the definition of the 12-hour time
          period).  If time remains in the 12-hour time period after meeting
          the technical acceptance criteria for the initial calibration,
          samples may be analyzed.  It is not necessary to analyze a
          continuing calibration standard if the initial calibration 'standard
          that is the same concentration as the continuing calibration
          standard meets the continuing calibration technical acceptance
          criteria.  A method blank is required.  Quantify all sample results
          against the initial calibration standard that is the same
          concentration as the continuing calibration standard (50 ug/L).

9.4.2.2   If time does not remain in the 12-hour period beginning with the
          injection of the instrument performance check solution, a new
          injection of the instrument performance check solution must be made.
          If the new injection meets the ion abundance criteria for BFB, then
          a continuing calibration standard may be injected.

9.4.3   Procedure for Continuing Calibration

                                   D-26/VOA                            OLM04.2

-------
                                               Exhibit  D  Volatiles  --  Section 9
                                                Calibration and Standardization
                                                         Continuing Calibration

•9.4.3.1    Set  up the  purge and trap  GC/MS  system per the requirements in
           Section 9.1.1.

 9.4.3.2    Add  10 uL of internal standard solution (Section 7.2.4.3) to the  5
           mL syringe  or volumetric flask containing the  continuing calibration
           standard (Section 7.2.4.6).  Analyze the  continuing calibration
           standard according to Section  10.

 9.4.4    Calculations  for Continuing  Calibration

 9.4.4.1    Calculate a relative response  factor (RRF) for each target  and
           system monitoring compound using Equation 1.

 9.4.4.2    Calculate the percent difference between  the continuing  calibration
           relative response factor and the most recent initial calibration
           mean relative response factor  for each purgeable target  and system
           monitoring  compound-using  Equation 3.

           EQ.   3
                                       RRF_ - RRF.
                         %Difference = 	==	 x 100
                                          RRF±


           Where,

           RRFC    =  Relative response factor from continuing calibration
                      standard
           RKF±     =  Mean relative, response factor from the most recent
                      initial calibration meeting technical acceptance criteria

 9.4.5   Technical Acceptance Criteria for Continuing Calibration

 9.4.5.1   The continuing calibration standard must be analyzed at the
           frequency described in Section 9.4.2 on a GC/MS system meeting the
           BFB and initial calibration technical acceptance criteria.

 9.4.5.2   The relative response factor (RRF)  for each purgeable target and
           system monitoring compound listed in Table 5 must be greater than or
           equal to the compound's minimum acceptable response factor listed in
         « Table 5.

 9.4.5.3   The relative response factor percent difference for each purgeable
           target and system monitoring compound listed in Table 5 must be less
           than or equal to the value listed.

 9.4.5.4   Up to two compounds may fail the requirements listed in Sections
           9.4.5.2 and 9.4.5.3 and still meet the minimum relative response
           factor criteria and percent difference criteria.  However, these
           compounds must have a minimum relative response factor greater than
           or equal to 0.010 and the percent difference must be within the
           inclusive range of ±40.0 percent.

 9.4.5.5   Excluding those ions in the solvent front, no quantitation ion may

                                    D-27/VOA                            OLM04.2

-------
Exhibit D Volatiles -- Section 3
Calibration and Standardization
Continuing Calibration

          saturate the detector.  Consult the manufacturer's instrument
          operating manual to determine how saturation is indicated for your
          instrument.

9.4.6   Corrective Action for Continuing Calibration

9.4.6.1   If the continuing calibration technical acceptance criteria are not
          met, recalibrate the GC/MS instrument according to Section 9.3.3.
          It may be necessary to clean the ion source, change the column, or
          take other corrective actions to achieve the continuing calibration
          technical acceptance criteria.

9.4.6.2   Continuing calibration technical acceptance criteria must be met
          before any samples, which include MS/MSD samples, or required blanks
          are analyzed.  Any samples or required blanks analyzed when
          continuing calibration technical acceptance criteria have not been
          met will require reanalysis at no additional cost to the Agency.
                                   •D-28/VOA                            OLM04.2

-------
                                             Exhibit D Volatiles -- Section 10
                                                                     Procedure
                                                            Sample Preparation

10.0 PROCEDURE

10.1 Sample Preparation

10.1.1  If insufficient sample amount (less than 90%, of the required amount)
        is received to perform the analyses,  the Contractor shall contact SMO
        to apprise them of the problem.   SMO will contact the Region for
        instructions.  The Region will either require that no sample analyses
        be performed or will require that a reduced volume be used for the
        sample analysis.  No other changes in the analyses will be permitted.
        The Contractor shall document the Region's decision in the SDG
        Narrative.

10.1.2  If multiphase samples  (e.g., two-phase liquid sample, oily
        sludge/sandy soil sample) are received by the Contractor, the
        Contractor shall contact SMO to apprise them of the type of sample
        received.  SMO will contact the Region.  If all phases of the sample
        are amenable to analysis, the Region may require the Contractor to do
        any of the following:

        •  Mix the sample and analyze an  aliquot from  the homogenized sample.

        •  Separate  the phases of the sample and analyze each phase
           separately.  SMO will provide  EPA sample  numbers  for  the
           additional phases,  if required.

        •  Separate  the phases,  and analyze one or more of  the phases, but
           not all of the phases.   SMO  will provide  EPA sample numbers for
           the additional phases, if  required.

        •  Do not analyze the sample.

10.1.2.1   If all of the phases are not amenable to  analysis (i.e.,  outside
           scope), the  Region may require the  Contractor to do any of the
           following:

           •  Separate  the phases and analyze  the phase(s) that  is amenable to'
              analysis.  SMO  will provide EPA  sample numbers for the
              additional phases, if required.

           •  Do not analyze  the sample.

10.1.2.2   No other  changes  in the  analyses will be  permitted.   The
           Contractor shall  document the  Region's decision  in  the  SDG
           Narrative.

10.1.3  Water Samples

10.1.3.1   All water samples  must be allowed  to warm to ambient  temperature
           before analysis.

10.1.3.2   Prior to  the analysis of samples,  establish the  appropriate purge
           and  trap  GC/MS  operating conditions,  as  outlined in Section  9.1,
           analyze  the  instrument performance  check solution (Section  9.2),

                                    D-29/VOA                            OLM04.2

-------
Exhibit D Volatiles -- Section 10
Procedure
Sample Preparation

           and calibrate the GC/MS system according to Sections 9.3 through
           9.4.6.

10.1.3.3   If time remains in the 12-hour period  (as described in Section
           9.3.2), samples may be analyzed without analysis of a continuing
           calibration standard.

10.1.3.4   If time does not remain in the 12-hour period since the injection
           of the instrument performance check solution, both the instrument
           performance check solution and the continuing calibration standard
           must be analyzed before sample analysis may begin  (see Section
           9.4.2).

10.1.3.5   Adjust the purge gas  (helium) flow rate to 25-40 mL/minute.
           Variations from this  flow rate may be necessary to achieve better
           purging and collection efficiencies for some compounds,
           particularly chloromethane and bromoform.

10.1.3.6   Remove the plunger from a 5 mL syringe and attach a closed syringe
           valve.  Open the sample or standard bottle which has been allowed
           to come to ambient temperature, and carefully pour the sample into
           the syringe barrel to just short of overflowing.  Replace the
           syringe plunger and compress the sample.  Open the syringe valve
           and vent any residual air while adjusting the sample volume to 5
           mL.  This process of  taking ari aliquot destroys the validity of
           the sample for future analysis so, if there is only one VOA vial,
           the analyst must fill a second syringe at this time to protect
           against possible loss of sample integrity.  This second sample is
           maintained only until such time-as the analyst has determined that
           the first sample has  been analyzed properly.  Filling one 5 mL
           syringe would allow the use of only one syringe.  If an analysis
           is needed from the second 5 mL syringe, it must be performed
           within 24 hours.  Care must also be taken to prevent air from
           leaking into the syringe.

10.1.3.7   Add 10 uL of the system monitoring compound spiking solution
            (Section.7.2.4.1) and 10 uL of the internal standard spiking
           solution  (Section 7.2.4.3) through the valve bore of the syringe,
           then close the valve. The system monitoring compounds and
           internal standards may be mixed and added as a single spiking
           solution.  The addition of 10 uL of the system monitoring compound
           spiking solution to 5 mL of sample is equivalent to a
           concentration of 50 ug/L of each system monitoring compound.
10.1.3.8
Attach the syringe-syringe valve assembly to the syringe valve on
the purging device.  Open the syringe valves and inject the sample
into the purging chamber.
10.1.3.9    Close  both valves  and purge  the  sample  for  11.0  (±0.1) minutes at
            ambient  temperature.

10.1.3.10   At  the conclusion  of  the purge time,  attach the  trap  to the
            chromatograph,  adjust the  device to  the desorb mode,  and begin the
            gas chromatographic temperature  program.  -Concurrently, introduce

                                   D-30/VOA                           OLM04.2

-------
                                             Exhibit D Volatiles -- Section 10
                                                                     Procedure
                                                            Sample Preparation

           the trapped materials to the gas chromatographic column by rapidly
           heating the trap to 180 °C while backflushing the trap with an
           inert gas between 20 and 60 mL/minute for four minutes.

10.1.3.11  While the trap is being desorbed into the gas chromatograph, empty
           the purging chamber.  Wash the chamber with a minimum of two 5 mL
           flushes of reagent water to avoid carryover of target compounds.
           For samples containing large amounts of water-soluble materials,
           suspended solids, high-boiling compounds, or high purgeable
           levels, it- may be necessary to wash out the purging device with a
           detergent solution between analyses, rinse it with distilled
           water, and then dry it in an oven at 105 °C.

10.1.3.12  After desorbing the sample for four minutes, recondition the trap
           by returning the purge and trap device to the purge mode.  Wait 15
           seconds, then close the syringe valve on the purging device to
           begin gas flow through the trap.  The trap temperature  should be
           maintained at 180 °C.  Trap temperatures up to 220 °C may be
           employed.  However, the higher temperature will shorten the useful
           life of the trap.  After approximately seven minutes, turn off  the
           trap heater and open the syringe valve to stop the gas  flow
           through the trap.  When cool, the trap is ready for the next
           sample.

10.1.4  Low Level Soil/Sediment Samples

10.1.4.1   The Contractor must determine whether a soil/sediment sample
           should be analyzed by the low or medium method.  It is  the
           responsibility of the Contractor to analyze the sample  at the
           correct level.

10.1.4.2   Three approaches may be taken to determine whether'the  low  level
           or medium level method must be  followed.

           •  Assume the sample is low level and analyze a 5 g sample.

           •  Use the X factor calculated from the hexadecane screen (Appendix
              A) to determine the appropriate method for analysis.

           •  Use other EPA approved screening procedures, or an in-house
              laboratory screening procedure.  The procedure must  be
              documented and available for review during on-site, laboratory
              evaluation or when requested by the Technical Project Officer or
              Administrative Project Officer.

10.1.4.3   If  the  on column concentration  of  any  target  compound  exceeds  the
           initial  calibration range  from  the  analysis of'5 g  sample,  a
           smaller sample  size must be  analyzed.  However,  the  smallest
           sample  size permitted  is  0.5  g.   If  smaller than 0.5 g sample  size
           is  needed to prevent the on  column concentration of  target
           compounds  from exceeding  the  initial calibration range, the medium
           level method must be used.
 10.1.4.4    The low level soil/sediment method is based on a heated purge of a

                                    D-31/VOA                           OLM04.2

-------
Exhibit D Volatiies -- Section 10
Procedure
Sample Preparation

           soil/sediment sample mixed-with reagent water containing the
           system monitoring compounds and the internal standards.  Analyze
           all matrix spike/matrix spike duplicate samples, blanks, and
           standards under the same condition as the samples.

10.1.4.5   The procedure described below is  to be utilized for  volatile low
           level soil samples unless  the Modified SW-846 Method 5035  is
           specified at the time of sample scheduling.  If the  Modified SW-
           846 Method 5035 is to be utilized for low level soil samples, use
           the procedure outlined in  Appendix B and follow the  equipment
           manufacturer's instructions.  Note: The requirements of this SOW
           must be met at all times.  If the above method is  specified at  the
           time of sample scheduling, but the Contractor believes  that the
           samples cannot be processed by this method, the Contractor shall
           immediately contact SMO.   SMO will contact  the Region for
           instructions.

10.1.4.6   Use 5 grams of sample, or  use the X Factor  (Appendix A) or your
           in-house screening procedure to determine the sample size  for
           purging.

10.1.4.7   Prior to the analysis of samples, establish the appropriate purge
           and trap GC/MS operating conditions, as outlined in  Section 9.1,
           analyze the instrument performance check solution  (Section 9.2),
           and calibrate the GC/MS system according to Sections 9.3 through
           9.4.6.  This should be done prior to the preparation of the sample
           to .avoid lose of volatiles from standards and sample.   A heated
           purge calibration curve must be prepared and used  for the
           quantitation of all samples analyzed with the low  level method.
           Follow the initial and daily calibration instructions  (Sections
           9.3.3 and 9.4.3), but increase the purge temperature to 40 °C.

10.1.4.8   To prepare the reagent water containing the system monitoring
           compounds and the internal standards, remove the plunger from a 5
           mL "Luerlok" type syringe  equipped with a syringe  valve and fill
           until overflowing with reagent water.  Replace the plunger and
           compress the water to vent trapped air.  Adjust the  volume to 5
           mL.  Add 10 uL of the.system monitoring compound spiking solution
           and 10 uL of the internal  standard solution to the syringe through
           the valve.  NOTE:  Up to 10 mL of reagent water may  be  added to a
           soil sample to increase purge gas/sample interaction.   All soil
           samples including MS and HSD, standards, and blanks  within an SDG
           must have the same amount  of reagent water  added.  Do not
           increase/change the amount of system monitoring compound and
           internal standard solution added.

10.1.4-. 9   The sample  (for volatile organics) is'defined as the entire
           contents of the sample container.  Do not discard  any supernatant
           liquids.  Mix the contents of the sample container with a  narrow
           metal spatula.  Weigh 5 g  or the  amount determined in using the
           screening procedure in Appendix A or an in-house screening
           procedure into a tared purge device.  Use a top loading balance.
           Note and record the actual weight to the nearest 0.1 g.


                                   D-32/VOA                            OLM04.2

-------
                                             Exhibit D Volatiles -- Section 10
                                                                     Procedure
                                                            Sample Preparation

10.1.4-.10  Add the spiked reagent water to the purge device and connect the
           device to the purge and trap system. .

10.1.4.11  NOTE:  Prior to the attachment of the purge device, the steps in
           Sections 10.1.4.8 and 10.1.4.10 above must be performed rapidly to
           avoid loss of volatile organics.  These steps must be performed in
           a laboratory free of solvent fumes.

10.1.4.12  Heat the sample to 40 °C  (± 1 °C)  and purge the sample for 11.0 (±
           0.1) minutes.

10.1.4.13  Proceed with the analysis as outlined in Sections 10.1.3.9 through
           10.1.3.12.

10.1.5  Medium Level Soil/Sediment Samples

10.1.5.1   The medium level soil/sediment method is based on extracting the
           soil/sediment sample with methanol.  An aliquot of the methanol
           extract is .added to reagent water containing the system monitoring
           compounds and the internal standards.  The reagent water
           containing the methanol extract is purged at ambient temperature.
           When using the screening method in Appendix A, all samples with an
           X Factor  > 1.0 should be  analyzed by the medium level method.

10.1.5.2   Prior to  the analysis of  samples, establish the appropriate purge
           and trap  GC/MS operating  conditions, as outlined in  Section 9.1,
           analyze the  instrument performance check solution  (Section 9.2),
           and calibrate the GC/MS system according to Sections 9.3  through
           9.4.6.  This should be done prior to the addition of the  methanol
           extract to reagent water.  Because the methanol extract and
           reagent water mixture is  purged at ambient temperature, the
           instrument performance check, initial calibration, and  continuing
           calibration  for water samples may be used  for analyses  of medium
           level soil/sediment sample extracts.

10.1.5.3   The  sample  (for volatile  organics)  is defined as the entire
           contents  of  the  sample container.  Do not  discard  any  supernatant
           liquids.  Mix the contents of the  sample container with a narrow
           metal spatula.  Weigh 4 g (wet weight)  into a tared  15  mL vial.
           Use  a top loading balance.  Record  the  actual weight to the
           nearest  0.1  g.

           NOTE: If methanol preserved sample is  to be analyzed,  weigh sample
           vial  and contents  to  the  nearest  0.1 g and record  the  weight.
           Record  any discrepancies  between  laboratory determined weight  and
           sampler determined  weight in  the  SDG Narrative  and utilize  the
           sampler determined  weight in  any  calculations.  Proceed to Section
           10.1.5.6.

10.1.5.4   Quickly add 10  mL of methanol  to  the vial.  Cap and  shake for  2
           minutes.

10..1.5.5   NOTE:   The  steps  in Sections  10.1.5.3  and  10.1.5.4 must be
           performed rapidly to  avoid loss of  volatile organics.   These steps

                                   D-33/VOA                            OLM04.2

-------
Exhibit D Volatiles -- Section 10
Procedure
Sample Preparation

           must be performed  in a  laboratory free of solvent fumes.

10.1.5.6   Let the solution settle.  Then, using a disposable pipette,
           transfer approximately  1 mL of extract into a GC vial for storage.
           The remainder may  be discarded.  The 1 mL extract may be stored in
           the dark at 4  °C  (±2  °C) prior to the analysis.

10.1.5.7   Table  6 can be used to  determine the volume of methanol extract to
           add to the 5 mL of reagent water for analysis.  If the hexadecane
           screen procedure  (Appendix A) was followed, use the estimated
           concentration  (Option A) or the X Factor  (Option B) to determine
           the appropriate volume.  Otherwise, estimate the concentration
           range  of the sample from the  low level analysis or from the in-
           house  screening procedure to  determine the appropriate volume.


10.1.5.8   Remove the plunger from a 5 mL "Luerlok" type syringe equipped
           with a syringe valve and fill until overflowing with reagent
           water.  Replace the plunger and compress the water to vent trapped
           air.   Adjust the volume to 4.9 mL.  Pull the plunger back to 5 mL
           to allow volume for the addition of sample and standards.  Add 10
           uL of  system monitoring compound and 10 uL of the internal
           standard solution. Also add  the volume of methanol extract
           determined in Section 10.1.5.7 and a volume of clean methanol to
           total  100 uL  (excluding methanol in system monitoring/internal
           standard solutions).

10.1.5.9   Attach the syringe-syringe valve assembly to the syringe valve on
           the purge device.  Open the syringe valve and inject the
           water/methanol sample into the purging chamber.

10.1.5.10  Proceed with the analysis as  outlined in Section 10.1.3.9 through
           10.1.3.12.

10.1.6  Sample Dilutions

10.1.6.1   If the on-column concentration of any target compound in any
           sample exceeds the initial calibration range, a new aliquot of
           that sample must be diluted and purged.  Guidance in performing
           dilutions and exceptions to this requirement are given in Sections
           10.1.6.2 through  10.1.6.10.

10.1.6.2   Use the results of the  original analysis to determine.the
           approximate dilution  factor required to get the largest analyte
           peak within the initial calibration range.

10.1.6.3   The dilution factor chosen should keep the response of the largest
           analyte peak for a target compound in the upper half of the
           initial calibration range of  the instrument.

10.1.6.4   For medium level  soil/sediment analyses, the purgeable organics
           screening procedure  (Appendix A), if used, will show the
           approximate concentrations of major sample components.  If a
           dilution of the sample  was -indicated, this dilution shall be made

                                   D-34/VOA                            OLM04.2

-------
                                             Exhibit D Volatiles -- Section 10
                                                                     Procedure
                                             pH Determination/Percent Moisture

           just prior to GC/MS analysis of the sample.  All steps in the
           dilution procedure must be performed without delays until the
           point at which the diluted sample is in a gas tight syringe.

10.1.6.5   For water samples, all dilutions are made in volumetric flasks  (10
           mL to 100 mL).  Select the volumetric flask that will allow for
           the necessary dilution.  Intermediate dilutions may be necessary
           for extremely large dilutions.  Calculate the approximate volume
           of reagent water  which will be added to the volumetric flask
           selected and add  slightly less than this quantity of reagent water
           to the flask.

10.1.6.6  . For water samples, inject the proper aliquot from the syringe
           prepared in Section 10.1.3.6 into the volumetric flask.  Only
           aliquots of 1 mL  increments are permitted.  Dilute the aliquot to
           the mark on the flask with reagent water.  Cap the flask, invert,
           and shake three times.

10.1.6.7   Fill a 5 mL syringe with the diluted sample as in Section
           10.1.3.6.

10.1.6.8   If this is an intermediate dilution, use it and repeat the above
           procedure to  achieve larger dilutions.

10.1.6.9   Do not submit data for more than two analyses, i.e., from the
           original sample and one dilution, or, if the volatile screening
           procedure was employed, from the most concentrated dilution
           analyzed.and  one  further dilution.

10.1.6.10  For total xylenes, where three isomers are quantified as two
           peaks, the calibration of each peak should be considered
           separately,  i.e., a diluted analysis is not required for total
           xylenes unless  the concentration of the peak representing the
           single isomer exceeds 200 ug/L  (ug/kg for  soils/sediment) or the
           peak  representing the two co-eluting isomers on the GC  column
           exceeds 400 ug/L  (ug/kg for soils/sediment).

10.2  pH Determination  (Water Samples)

      Once  the  sample aliquots have been  taken  from  the VOA vial, the pH of
      the water  sample must be determined.  The purpose of the pH
      determination  is to ensure  that all VOA samples  were acidified  in the
      field.   Test the pH by  placing one  or two drops  of sample  on  the pH
      paper (do  not  add  pH  paper  to the vial).  Record the pH of each  sample,
      and report  these data in the SDG Narrative, following the  instructions
      in Exhibit  B.  No  pH  adjustment is  to be  performed by the  Contractor.

10.3  Percent Moisture Determination

      Immediately after  weighing  the sample for analysis, weigh  5-10  g  of  the
      soil/sediment  into a  tared  crucible.  Determine  the percent moisture by
      drying  overnight  at 105 °C.  Allow  to cool in  a  desiccator before
      weighing.   Concentrations of  individual analytes will be reported
      relative to the  dry weight  of soil/sediment.

                                   D-35/VOA                            OLM04.2

-------
Exhibit D Volatiles -- Section 10
Procedure
pH Determination/Percent Moisture
     EQ.
                       grams of wet sample - grams of dry sample   „„„
           %Moisture =  —	  x 100
                                 grams of wet sample
                                   D-36/VOA                            OLM04.2

-------
                                             Exhibit D Volatiles -- Section 11
                                                Data Analysis and Calculations
                                                    Qualitative Identification
11.0  DATA ANALYSIS AND CALCULATIONS
11.1  Qualitative Identification

11.1.1  Identif icati'on of Target Compounds

11.1.1.1  The compounds listed in the Target Compound List (TCL) in Exhibit C
          (Volatiles) shall be identified by. an analyst competent in the
          interpretation of mass spectra  (see Exhibit A, Section 4.3.1) by
          comparison of the sample mass spectrum to the mass spectrum of a
          standard of the suspected compound.  Two criteria must be satisfied
          to verify the identifications:   (1) elution, of the sample component
          at the same GC relative retention time as the standard component,
          and  (2) correspondence of the sample component and standard
          component mass spectra.

11.1.1.2  For establishing correspondence of the GC relative retention time
          (RRT), the sample component RRT must compare within ±0.06 RRT units
          of the RRT of the standard component.  For reference, the standard
          must be run in the same 12-hour time period as the sample.  If
          samples are analyzed during the same 12-hour time period as the
          initial calibration standards, use the RRT values from the 50 ug/L
          standard.  If co-elution of interfering components prohibits
          accurate assignment of the sample component RRT from  the total ion
          chromatogram, the RRT should be assigned by using extracted ion
          current profiles for ions unique to the component of  interest.

11.1.1.3  For  comparison of standard and  sample component mass  spectra, mass
          spectra obtained on the Contractor's GC/MS are required.  Once
          obtained,  these standard spectra may be used for identification
          purposes,  only if the Contractor's GC/MS meets the daily instrument
          performance requirements for BFB.  These standard spectra may be
          obtained from the run used to obtain reference RRTs.

11.1.1.4  The  requirements for qualitative verification by comparison of mass
          spectra are as follows:

          •    All ions present in the standard mass spectra at  a relative
               intensity greater than 10.0 percent  (most abundant ion in the
               spectrum equals 100.0 percent) must be present  in the sample
               spectrum.

          •    The relative intensities of ions specified above  must agree
               within ± 20.0 percent between  the standard and  sample spectra.
               (Example:  For an  ion with  an  abundance of 50.0 percent  in the
               standard spectra,  the corresponding sample abundance must be
               between 30.0 and 70.0 percent).

          •    Ions greater than  10.0 percent  in the sample spectrum but not
               present in the standard spectrum must be  considered and
               accounted  for by the analyst making the comparison.  In  Exhibit
               A,  Task II, the verification process should favor false
               positives.  All compounds meeting the identification criteria
               must be reported with their spectra.  For all compounds  below
               the CRQL,  report the actual value followed by a "J", e.g.,  "3J".
                                    D-37/VOA
OLM04.2

-------
Exhibit D Volatiles -- Section 11
Data Analysis and Calculations
Qualitative Identification

11.1.1.5  If a compound cannot be verified by all of the criteria in Section
          11.1.1.4, but in the technical judgment of the mass spectral
          interpretation specialist, the identification is correct, then the
          Contractor shall report that identification and proceed with
          guantitation in Section 11.2.

11.1.2  Identification of Non-Target Compounds

11.1.2.1  A library search shall be executed for non-target sample components
          for the purpose of tentative identification.  For this purpose, the
          NIST/EPA/NIH  (May 1992 release or later) and/or Wiley.  (1991 release
          or later), or equivalent mass spectral library, shall be used.

11.1.2.2. Up to 30 organic compounds of greatest apparent concentration not
          listed in Exhibit C for the volatile and semivolatile organic
          fraction, excluding the system monitoring compounds and internal
          standard compounds, shall be tentatively identified via a forward
          search of the NIST/EPA/NIH  (May 1992 release or later) and/or Wiley
          (1991 release or later), or equivalent mass spectral library.  The
          following are not to be reported:  1) Substances with responses less
          than 10' percent' of the internal standard (as determined by
          inspection of the peak areas or height), 2)  Substances which elute
          earlier than 30 seconds before the first purgeable compound listed
          in Exhibit C  (Volatiles) or three minutes after the last purgeable
          compound listed in Exhibit C  (Volatiles) has eluted are not required
          to be searched in this fashion, 3) Carbon dioxide, and 4)
          Semivolatile TCL compounds listed in Exhibit C.  Only after visual
          comparison of sample spectra with the nearest library searches will
          the mass spectral interpretation specialist assign a tentative
          identification.  NOTE:  Computer generated library search routines
          must not use normalizations which would misrepresent the library or
          unknown spectra when compared to each other.         . -

11.1.2.3  Up to 20 peaks of greatest apparent concentration  (as determined by-
          inspection of peak areas or heicjhts) that are suspected to be
          straight-chain, branched, or cyclic alkanes, alone or part of an
          alkane series shall be library searched.  Documentation for the
          tentative identification must be supplied.  Alkanes are not counted
          as part of the 30 organic compounds described in Section 11.1.2.2.

11.1.2.4  Guidelines for making tentative identification:

          •   Relative intensities of major ions in the reference spectrum
               (ions greater than 10.0 percent of the most abundant ion) should
              be present in the sample spectrum.

          •   The relative intensities of the major ions should agree within
              ±20.0 percent.  Example:  For an ion with an abundance of 50.0
              percent of the standard spectra, the corresponding sample ion
              abundance must be between 30.0 and 70.0 percent.

          •   Molecular ions present in reference spectrum should be present
              in sample spectrum.

          •   Ions present in the sample spectrum but not in the reference

                                   •D-3 8/VGA                            OLM04.2

-------
                                             Exhibit D Volatiles --  Section 11
                                                Data Analysis and Calculations
                                                                  Calculations

              spectrum should be reviewed for possible background
              contamination or presence of co-eluting compounds.

          •   Ions present in the reference spectrum but not in the  sample
              spectrum should be reviewed for possible subtraction from the
              sample spectrum because of background contamination or co-
              eluting compounds.  Data system library reduction programs can
              sometimes create these discrepancies.

11.1.2.5  If, after careful review and in the technical judgement of the mass
          spectral interpretation specialist, no valid identification can be
          made, the compound should be reported as follows:

          •   If the library search produces a match at or above 85%, 'report
              that compound.

          •   If the library search produces more •;than one compound at or
              above 85%, report the first compound  (highest).

          •   Tf the library search produces no matches at or above 85%, the
              compound should be reported as unknown.  The mass spectral
              specialist should give additional classification of the unknown
              compound, if possible  (i.e. unknown aromatic, unknown
              hydrocarbon, unknown acid type, unknown chlorinated compound).
              If probable molecular weights can be distinguished, include
              them.

11.1.2.6  Straight-chain, branched, or cyclic alkanes are not to be reported
          as tentatively identified compounds on FORM I VOA-TIC.  When the
          above alkanes are tentatively identified, the concentration (s) are
          to be estimated as described in Section 11.2.2 and reported in the
          SPG Narrative as alkanes, by class  (i.e., straight-chain,  branched,
          or cyclic; as a series; as applicable).

11.2  Calculations

11.2.1  Target Compounds

11.2.1.1  Target compounds identified shall be quantified by the internal
          standard method using the equations below.  The internal standard
          used shall be that which is assigned in Table 3.  The relative
          response factor  (RRF) from the continuing calibration standard is
          used to calculate the concentration in the sample.

11.2.1.2  Water

          EQ.'.5

                                            (AJ  (I)  (Df)
                    Concentration (ug/L)  =
                                           (A.s)  (RRF)  (Vo)
          Where,
                                    D-39/VOA                           OLM04.2

-------
  Exhibit D Volatiles -- Section 11
  Data Analysis and Calculations
  Calculations

            A^      =  Area of the characteristic ion  (EICP) for the compound to
                       be measured (see Table 2)
            Ais      =  Area of the characteristic ion  (EICP) for the specific
                       internal standard  (see Tables 3 and 4)
            Ia      =  Amount of internal standard added in nanograms  (ng)
            RRF     =  Relative response factor from the ambient temperature
                       purge of the calibration standard.
            V0      =  Volume of water purged in milliliters  (mL)
            Df      =  Dilution factor.  The dilution factor for analysis of
                       water samples for volatiles by this method is defined as
                       the ratio of the number of milliliters  (mL) of water
                       purged (i.e.,  V0 above)  to the number of mL of the
                       original water sample used for purging.  For example, if
                       2.0 mL of sample is diluted to 5 mL with reagent water
                       and purged, Df = 5 mL/2.0 mL = 2.5.
                       If no dilution is performed, Df = 1.

|  11.2.1.3  Low Level Soil/Sediment

            EQ.  6

                   Concentration (ug/Kg).  (dry weight basis) =
                                                              (Ais)  (RRF)  (W8)  (D)
          Where,

            Ax,  Is, Ais are as given for water, Equation 5.
            RRF =   Relative response factor from the heated purge of the
                    calibration standard.
                      100 - %moisture
            D   = '
            Ws  =   Weight of sample added to the purge tube, in grams  (g).

|  11.2.1.4  Medium Level Soil/Sediment

            EQ.  7
                                                           (A ) (I )  (V )(1000) (Df)
                  Concentration ug/Kg (dry weight basis) =
                                                            (Ais)  (RRF) (Va) (Ws) (D)
            Where,

            Ax,  Is, Als are  as  given for water,  Equation 5.
            RRF =   Relative  response  factor from the  ambient temperature purge
                    of the calibration standard.
            Vc  =   Total volume of the methanol extract in milliliters  (mL).
                    NOTE:  This volume is typically 10 mL, even  though only  1 mL
                    is transferred to  the vial  in Section 10.1.5.6.
            Va  =   Volume of the aliquot of the sample methanol extract  (i.e.,
                    sample extract not including the methanol added  to equal 100
                    uL) in microliters (uL) added to reagent water for purging.
            Ws  =   Weight of soil/sediment extracted, in grams  (g).

                                     D-40/VOA                            OLM04.2

-------
                                             Exhibit  D Volatiles --  Section 11
                                                Data  Analysis and Calculations
                                                                  Calculations

                   100 - %moisture
          D   =          100
          Df  =   Dilution factor.  The dilution factor for analysis of
                  soil/sediment samples for volatiles by the medium level
                  method is defined as:
                uL most cone, extract used to make dilution + uL clean solvent
                         uL most cone,  extract used to make dilution
11.2.1.5  For water,  low level and medium level soil/sediment samples,  xylenes
          (o-,m-, and p-isomers)  are to be reported as xylenes (total).
          Because the o- and p-xylene isomers co-elute on packed columns,  and
          the m- and p-xylene isomers co-elute on capillary columns, special
          attention must be given to the quantitation of the xylenes.  The
          relative response factor  (RRF) determined in Section 9.4.4 is based
          on the peak that represents the single isomer on the GC column used
          (m-xylene on packed columns, o-xylene on capillary columns).   In
          quantitating sample concentrations, use the areas on both peaks and
          the RRF from Section 9.4.4.  The areas of the two peaks may be
          summed, and the concentration determined, or the concentration
          represented by each of the two peaks may be determined separately,
          and then summed.  It is required that all three xylene isomers be
          present in the initial and continuing calibration standards.

11.2.1.6  The cis and trans stereo  isomers of 1,2-dichloroethene are to be
          reported separately.

11.2.1.7  Secondary ion quantitation is allowed only when there are sample
          matrix interferences with the primary ion.  If secondary ion
          quantitation is performed, document the reasons in the SDG
          Narrative.  A secondary ion cannot be used unless a relative
          response factor is calculated using the secondary ion.

11.2.1.8  The requirements listed in Sections 11.2.1.9 and 11.2.1.10 apply to
          all standards, samples including MS/MSDs, and blanks.

11.2.1.9  It is expected that situations will arise where the automated
          quantitation procedures in the GC/MS software provide inappropriate
          quantitations.  This normally occurs when there is compound co-
          elution, baseline noise,  or matrix interferences.  In these
          circumstances the Contractor must perform a manual quantitation.
          Manual quantitations are  performed by integrating the area of the
          quantitation ion of the compound.  This integration shall only
          include the area attributable to the specific TCL compound.  The
          area integrated shall not include baseline background noise.  The
          area integrated shall not extend past the point where the sides of
          the peak intersect with the baseline noise.  Manual integration is
          not to be used solely to  meet QC criteria, nor is it to be used as a

                                    D-41/VGA                            OLM04.2

-------
Exhibit D Volatiles -- Section Ii
Data Analysis and Calculations
Calculations

          substitute for corrective action on the chromatographic system.  Any
          instances of manual integration must be documented in the SDG
          Narrative.

11.2.1.10     In all instances where the data system report has been edited,
              or where manual integration or quantitation has been performed,
              the GC/MS operator must identify such edits or manual procedures
              by initialing and dating the changes made to the report, and
              shall include the integration scan range.  In addition, a
              hardcopy printout of the EICP of the quantitation ion displaying
              the manual integration shall be included in the raw data.  This
              applies to all compounds listed in Exhibit C (Volatiles),
              internal standards, and system monitoring compounds.

11.2.2  Non-Target Compounds

11.2.2.1  An estimated concentration for non-target compounds tentatively
          identified shall be determined by the internal standard method.  For
          quantitation, the nearest internal standard free of interferences
          shall be used.

11.2.2.2  The formulas for calculating concentrations are the same as in
          Sections 11.2.1.2, 11.2.1.3, and 11.2.1.4.  Total area counts  (or
          peak heights) from the total ion chromatograms are to be used for
          both the compound to be measured and the internal standard.  A
          relative response factor  (RRF) of one (1) is to be assumed.  The
          resulting concentration shall be qualified as "J"  (estimated, due to
          lack of a compound-specific response factor), and "N"  (presumptive
          evidence of presence), indicating the quantitative and qualitative
          uncertainties associated with this non-target component.  An
          estimated concentration must be calculated for all tentatively
          identified compounds as well as those identified as unknowns.

11.2.3  CRQL Calculations

        NOTE:  If the adjusted CRQL is less than the CRQL listed in Exhibit C
         (Volatiles), report the CRQL listed in Exhibit C  (Volatiles).

11.2.3.1  Water

          EQ.  8

                                                   Vx
                    Adjusted CRQL = Contract CRQL x —- x Df
          Where,

          V0 and Df are as given in Equation 5
          Vx = Contract Sample Volume (5 mL)
                                   D-42/VOA      '                      OLM04.2

-------
                                             Exhibit D Volatiles -- Section 11
                                                Data Analysis and Calculations
                                                                  Calculations
11.2.3.2  Low Level Soil/Sediment

          EQ.  9
                                                     (Wx)
                   Adjusted CRQL = Contract CRQL x
                                                   (WJ (D)
          Where,

          Ws and  D are as given in Equation 6
          Wx =    Contract Sample Weight  (5 g)

11.2.3.3  Medium Level Soil/Sediment

          EQ.  10

            Adjusted CRQL = Contract CRQL x
(Wx) (Ve) (Vy) (1000) (Df)
    (W.) (V) (V.)  (D)
          Where,

          Vt,  Df,  Ws, Va and D are as given in Equation 7
          Wx =    Contract Sample Weight  (4 g)
          Vy =    Contract Soil Aliquot Volume from soil methanol  extract
                   (100 uL)                             -
                  Contract Soil Methanol  Extract Volume  (10,000  uL)

11.2.4  System Monitoring Compound Recoveries

11.2.4.1  Calculate the recovery of  each  system monitoring  compound  in  all
          samples, blanks, matrix spikes, and matrix  spike  duplicates.
          Determine if the recovery  is within limits  (see Table  7),  and report
          on  the  appropriate  form.

11.2.4.2  Calculate the concentrations of the system  monitoring  compounds
          using the same equations as used  for target compounds.

11.2.4.3  Calculate the recovery of  each  system monitoring  compound  as
          follows:
           EQ.   11
                ._           Concentration (amount)  found
                %Recovery = 	 x 100
                            Concentration (amount) spiked
 11.2.5   Internal  Standard Responses and Retention Times

         Internal  standard responses and retention times  in all samples must be
         evaluated during or immediately after data acquisition.   Compare the
         sample internal  standard responses and retention times to the
         continuing calibration internal standard response and retention times.

        •                            D-43/VOA                            OLM04.2

-------
 Exhibit  D Volatiles --  Section II
 Data Analysis  and Calculations
.Technical Acceptance Criteria for Sample Analysis

         For  samples analyzed during the same 12-hour time period as  the
         initial calibration standards,  compare the internal standard responses
         and  retention times against the 50 ug/L calibration standard.   The
         extracted ion current profile'(EICP)  of  the internal standards  must  be
         monitored and evaluated for each sample,  blank,  matrix spike, and
         matrix spike duplicate.

 11.3   Technical Acceptance Criteria for Sample Analysis

 11.3.1  The  samples must be analyzed on a GC/MS  system meeting the BFB,
         initial calibration,  continuing calibration,  and blank technical
         acceptance criteria.

 11.3.2  The  sample must be analyzed within the contract  holding time.

 11.3.3  The  sample must have an associated method blank  meeting the  blank
         technical acceptance criteria.

 11.3.4  The  percent recovery of each of the system monitoring compounds in the
         sample must be within the acceptance windows in  Table 7.

 11.3.5  The  EICP area for each of the internal standards must be within the
         inclusive range of -50.0 percent and +100.0 percent of the response  of
         the  internal standards in the most recent continuing calibration
         analysis.

 11.3.6  The  retention time shift for each of the internal standards  must  be
         within ± 0.50 minutes  (30 seconds) between the sample and the most
         recent continuing calibration standard analysis.

 11.3.7  The  relative retention time (RRT) of the system  monitoring compound  in
         a  sample must be within ±0.06 (RRT) units of its relative retention
         time in the continuing calibration standard.

 11.3.8  Excluding those ions in the solvent front, no ion may saturate  the
         detector.  No target compound concentration may  exceed the upper limit
         of  the initial calibration range unless a more diluted aliquot  of the
         sample is also analyzed according to the procedures in Section  10.1.6.

 11.3.9  The  Contractor must demonstrate that there is no carryover from a
         contaminated sample before data from subsequent  analyses may be
         submitted.  After a sample that contains a target compound at a level
         exceeding the initial calibration range,  the Contractor must either:

         •      Analyze an instrument blank immediately after the contaminated
               sample.  If an autosampler is used, an instrument blank must
               also be analyzed using the same purge inlet that was used for
               the contaminated sample.   The instrument blanks must meet the
               technical acceptance criteria for blank analysis (see  Section
               12.1.4),  or

         •      Monitor the sample analyzed immediately after the contaminated
               sample for all compounds that were in the  contaminated sample
               and that exceeded the calibration range.  The maximum
               contamination criteria are as follows:   the sample must not

                                    D-44/VOA                            OLM04.2

-------
                                             Exhibit D Volatiles -- Section 11
                                                Data Analysis and Calculations
                                         Corrective Action for Sample Analysis

              contain a concentration above the CRQL for the target compounds
              that exceeded the limits in the contaminated sample.  If an auto
              sampler is used,  the next sample analyzed using the same purge
              inlet that was used for the contaminated sample must also meet
              the maximum contamination criteria.   If the maximum criteria
              were exceeded, then all samples affected by the carryover must
              be reanalyzed at no additional cost to the Agency.

11.4  Corrective Action for Sample Analysis

11.4.1  Sample technical acceptance criteria must be met before data are
        reported.  Samples contaminated from laboratory sources or sample
        results which failed to meet the sample technical acceptance criteria
        require reanalysis at no additional cost to the Agency.

11.4.2  Corrective actions for failure to.meet instrument performance checks,
        initial and continuing calibration, and method blanks must be
        completed before the analysis of samples.

11.4.3  Corrective action for system monitoring compounds and internal
        standard compounds that fail to meet acceptance criteria.

11.4.3.1  If any of the system monitoring compounds and internal standard
          compounds fail to meet acceptance criteria:

          •   Check all calculations, instrument logs, the system monitoring
              compound and internal standard compound spiking solutions, and
              the instrument operation.  If the calculations were incorrect.,
              correct the calculations and verify that the system monitoring.
              compound recoveries and internal standard compound responses
              meet acceptance criteria.

          •   If the instrument logs indicate that the incorrect amount of
              system monitoring compound or internal standard compound spiking
              solution was added, then reanalyze the sample after adding the
              correct amount of system monitoring compound and  internal
              standard spiking solutions.

          •   If the system monitoring compound spiking solution or  internal
              standard compound spiking solution was improperly prepared,
              concentrated, or degraded, re-prepare the solutions and
              reanalyze the samples.

          •   If the instrument malfunctioned, correct the instrument problem
              and reanalyze the sample.  If the instrument malfunction
              affected the  calibration, recalibrate the instrument before
              reanalyzing the sample.  Verify that the system monitoring
              compound recoveries meet acceptance criteria.

11.4.3.2  If  the above actions do not correct the problem, then the  problem
          may be due to a sample matrix effect.  To determine  if  there was  a
          matrix effect, take the following corrective action  steps:

          •   Reanalyze the sample.  EXCEPTION:  If system monitoring compound
              recoveries or internal standard compound response in a sample

                                    D-45/VOA                            OLM04.2

-------
Exhibit D Voiatiies -- Section II
Data Analysis and Calculations
Corrective Action for Sample Analysis

              used for a matrix spike or matrix spike duplicate were outside
              the acceptance criteria, then it should be reanalyzed only if
              system monitoring compound recoveries and internal standard
              compound response met acceptance criteria in both the matrix
              spike and matrix spike duplicate analyses.

          •   If the system monitoring compound recoveries and the internal
              standard compound responses meet the acceptance criteria in the
              reanalyzed sample, then the problem was within the Contractor's
              control.  Therefore, submit data only from the reanalysis.

          •   If the system monitoring compound recoveries and/or the internal
              standard compound responses fail to meet the acceptance windows
              in the reanalysis, then submit data from both analyses.
              Distinguish between the initial analysis and the reanalysis on
              all deliverables using the suffixes on Exhibit B.

11.4.4  Corrective action for system monitoring compounds relative retention
        times/internal standard compounds retention times outside acceptance
        criteria.

11.4.4.1  If the system monitoring compounds relative retention times .or
          internal standard compounds retention times are not within their
          acceptance criteria, check the instrument for malfunctions.  If the
          instrument malfunctioned, correct the instrument problem and
          reanalyze the sample.  If the instrument malfunction affected the
          calibration, recalibrate the instrument before reanalyzing the
          samples.

11.4.4.2  If the above actions do not correct the problem, then the problem
          may be due to a sample matrix effect.  To determine if there was a
          matrix effect, take the following corrective action steps:

          •   Reanalyze the sample.  EXCEPTION:  If the system monitoring
              compounds relative retention times or internal standard
              compounds retention times in a sample used for a matrix spike or
              matrix spike duplicate were outside the acceptance criteria,
              then it should be reanalyzed only if the system monitoring
              compounds and internal standard compounds retention times were
              within the acceptance criteria in both the matrix spike and
              matrix spike duplicate analyses.

          •   If the system monitoring compounds relative retention times and
              internal standard compounds retention times are within the
              acceptance criteria, then the problem was within the
              Contractor's control.  Therefore, submit only data from the
              reanalysis when the system monitoring compounds relative
              retention times and the internal standard compounds retention
              times are within the acceptance limits.
                                   D-46/VOA                            OLM04.2

-------
                               Exhibit D Volatiles -- Section 11
                                  Data Analysis and Calculations
                           Corrective Action for Sample Analysis

If the system monitoring compounds relative retention times or
the internal standard compounds retention times are outside the
acceptance criteria in the reanalysis, then submit data from
both analyses.   Distinguish between the initial analysis and the
reanalysis on all deliverables, using the suffixes in Exhibit B.
                      D-47/VOA                            OLM04.2

-------
Exhibit D Volatiles -- Section 12
Quality Control
Blank Analyses

12.0  QUALITY CONTROL

12.1  Blank Analyses

12.1.1  Summary -- There are three different types of blanks required by this
        method.

12.1.1.1  METHOD BLANK - a volume of a clean reference matrix (reagent water
          for water samples or a purified solid matrix for soil/sediment
          samples) that is carried through the entire analytical procedure.
          The volume or weight of the reference matrix must be approximately
          equal to the volume or weight of samples associated with the blank.
          The purpose of a method blank is to determine the levels of
          contamination associated with the processing and analysis of
          samples.

12.1.1.2  STORAGE BLANK - upon receipt of the first samples in an SDG, two
          40.0 mL screw-cap volatile vials with a PTFE-faced silicone septum
          are filled with reagent water (80 mL total).  The vials are stored
          with the samples in the SDG under the same conditions.  After all
          samples in the SDG have been analyzed, the storage blank is
          analyzed.  The storage blank indicates whether contamination may
          have occurred during storage of samples.

12.1.1.3  INSTRUMENT BLANK - a 5.0 mL aliquot of reagent water that is
          carried through the entire analytical procedure.  Instrument blanks
          are analyzed after a sample/dilution which contains a target
          compound exceeding the initial calibration range1:  The results from
          the instrument blank analysis indicate whether there is
          contamination from a previous sample.

12.1.2  Frequency of Blank Analyses

12.1.2.1  The method blank must be analyzed at least once during every 12-
          hour time period on each GC/MS system used for volatile analysis
          (see Section 9.2.2 for the definition of the 12-hour time period).

12.1.2.2  The method blank must be analyzed after the continuing calibration
          and before any samples, including matrix spike/matrix spike
          duplicates, dilutions, or storage blanks are analyzed.  The method
          blank must be analyzed after the initial calibration sequence if
          samples are analyzed before the 12-hour period expires.  A method
          blank must be analyzed in each 12-hour time period in which samples,
          including dilutions, matrix spikes/matrix spike duplicates, and
          storage blanks are analyzed.

12.1.2.3  A minimum of one storage blank must be analyzed per SDG after all
          samples for that SDG have been analyzed.

12.1.2.4  The Contractor must demonstrate that there is no carryover from
          contaminated samples before data from subsequent analyses may be
          used.   Samples/dilutions may contain target compounds at levels
          exceeding the initial calibration range.  An instrument blank must
          be analyzed after the sample that exceeds the calibration range
          (also in the same purge inlet if an autosampler is used) or a sample

                                  . D-48/VOA                            OLM04.2

-------
                                             Exhibit  D  Volatiles  --  Section  12
                                                               Quality Control
                                                               Blank Analyses

          that meets  the  maximum contamination criteria in Section 11.3.8 must
          be  analyzed.  For these purposes,  if the  instrument blank meets the
          technical acceptance criteria for  blank analyses or the sample meets
          the maximum contamination criteria,  the system is considered to be
          uncontaminated.   If the instrument blank  or sample does not meet the
          criteria (i.e.,  contaminated),  the system must be decontaminated.
          Until  an instrument blank meets the blank technical acceptance
          criteria or a sample meets the maximum contamination criteria in
          Section 11.3.8;  any samples analyzed since  the original contaminated
          sample will require reanalysis at  no additional cost to the Agency.
          NOTE:   Only the instrument blank which demonstrates that there was
          no  carryover from the previous sample or  the  instrument blank that
          demonstrates that the system is clean (Section 12.1.4)  needs to be
          reported.   Instrument blanks analyzed during  the instrument
          decontamination process which exceed the  requirements listed in
          Section 12.1.4  do not need to be reported.

12.1.3  Procedure for Blank Analyses

12.1.3.1  For water samples, a volatile method blank  consists of  a 5 mL volume
          of  reagent  water  (Section 7.1.1) spiked with  10 uL of the system
          monitoring  compound spiking solution (Section 7.2.4.1)  and 10 uL  of
          the internal standard spiking solution (Section 7.2.4.3) and carried
          through the analytical procedure.

12.1.3.2  For low level soil/sediment samples, a volatile method blank
          consists of 5 g of a purified solid matrix  added to 5 mL of reagent
          water that  has  been spiked with 10 uL each  of the system monitoring
          compound spiking solution and the  internal  standard spiking
          solution.  This method blank is then carried  through the analytical
          procedure.

12.1.3.3  For medium level soil/sediment samples, a volatile method blank
          consists of 4 g of a purified solid matrix  added to 10 mL of
          methanol and extracted for two minutes.  A 100 uL aliquot of the
          methanol is added to reagent water and spiked with 10 uL of the
          internal standard spiking solution and 10 uL  of the system
          monitoring compound spiking solution and taken through the
          analytical procedure.

12.1.3.4  Storage/instrument blanks consist of a 5 mL volume of reagent water -
          (Section 7.1.1) spiked with 10 uL of the system monitoring compound
          spiking solution  (Section 7.2.4.1) and 10 uL.of the internal
          standard spiking  solution  (Section 7.2.4.3) and carried through the
          analytical procedure.

12.1.3.5  A  storage blank shall be analyzed and reported as a water sample
          unless  the SDG contains only soil samples.   If an SDG contains only
          soil samples, the storage blank may be analyzed and reported as a
          soil sample.

12.1.3.6  Identify and quantitate analytes'according to Section 11.0.

12.1.4  Technical Acceptance Criteria for Blank Analyses


                                   D-49/VOA                            OLM04.2

-------
Exhibit D Volatiles -- Section 12
Quality Control
Blank Analyses

12.1.4.1  All blanks must be analyzed on a GC/MS system meeting the BFB,
          initial calibration and continuing calibration technical acceptance
          criteria and at the frequency" described in Section 12.1.2.

12.1.4.2  A storage blank shall be analyzed and reported as a water sample
          unless the SDG contains only soil samples.  If an SDG contains only
          soil samples, the storage blank may be analyzed and reported as a
          soil sample.

12.1.4.3  The percent recovery of each of the system monitoring compounds in a
          blank must be within the acceptance windows in Table 7.

12.1.4.4  The EICP area for each of the internal standards in a blank must be
          within the inclusive range of -50.0 percent and +100.0 percent of
          the response of the internal standards in the most recent continuing
          calibration analysis.

12.1.4.5  The retention time shift for each of the internal standards in a
          blank must be within ±0.50 minutes (30 seconds) of its retention
          time in the most recent continuing calibration standard analysis.

12.1.4.6  The concentration of each target compound found in the blank must be
          less than its .CRQL listed in Exhibit C (Volatiles),  except for
          methylene chloride and cyclohexane which must be less than 2.5 times
          their respective CRQLs, and acetone and 2-butanone,  which must be
          less than 5 times their respective CRQLs.

12.1.5  Corrective Action for Blank Analyses

12.1.5.1  It is the Contractor's responsibility to ensure that method
          interferences caused by the contaminants in solvent, reagents,
          glassware, laboratory air, and other sample storage and processing
          hardware that lead to discrete artifacts and/or elevated baselines
          in gas chromatograms,  be eliminated.  If a Contractor's blanks
          exceed the criteria in Section 12.1.4.6,  the Contractor must
          consider the analytical system to be out of control.  The source of
          the contamination must be investigated and appropriate corrective
          measures must be taken and documented before further analysis
          proceeds.

12.1.5.2  Any method blank or instrument blank that fails to meet the
          technical acceptance criteria for blank analyses must be reanalyzed
          at no additional cost to the Agency.  Furthermore, all samples,
          including MS/MSD samples, processed within the 12-hour period with a
          method blank or instrument blank that does not meet the technical
          acceptance criteria for blanks will require reanalysis at no
          additional cost to the Agency.

12.1.5.3  If the storage blank does not meet the technical acceptance criteria
          for blank analyses in Sections 12.1.4.1 through 12.1.4.5, correct
          system problems and reanalyze the storage blank.  If the storage
          blank does not meet the criteria in Section 12.1.4.6, reanalyze the
          storage blank to determine whether the contamination occurred during
          storage or during the analysis.  If, upon reanalysis, the storage
          blank meets the criteria in Section 12.1.4.6, the problem occurred

                                   D-50/VOA                            OLM04.2

-------
                                             Exhibit D Volatiles -- Section 12
                                                               Quality Control
                                  Matrix Spike/Matrix Spike Duplicate (MS/MSD)

          during the analysis and the reanalyzed storage blank results must be
          reported.   If upon reanalysis,  the storage blank did not meet the
          criteria in Section 12.1.4.6,  the problem occurred during storage.
          The laboratory manager or his/her designee must address the problem
          in the SDG Narrative and discuss the corrective actions implemented
          to prevent future occurrences.   .NOTE:   A copy of the storage blank
          data must also be retained by the Contractor and be made available
          for inspection during on-site laboratory evaluations.

12.2  Matrix Spike/Matrix Spike Duplicate  (MS/MSD)

12.2.1  Summary of MS/MSD

        In order to evaluate the effects of the sample matrix on the methods
        used for volatile analyses, the Agency has prescribed a mixture of
        volatile target compounds to be spiked into two aliquots of a sample,
        and analyzed in accordance with the appropriate method.

12.2.2  Frequency of MS/MSD

12.2.2.1  A matrix spike and matrix spike duplicate must be performed for each
          group of samples of a similar matrix for the following, whichever is
          most frequent:

          •   Each SDG, or

          •   Each matrix within an SDG, or

          •   Each group of samples of a similar concentration level  (soils
              only).

          MS/MSD samples shall be analyzed unless otherwise specified on the
          Traffic Report  (TR).  If no MS/MSD samples are specified on the TR,
          the Contractor shall contact SMO to confirm that MS/MSD analyses are
          not required.

12.2.2.2  As a part of  the Agency's QA/QC program, water rinsate samples
          and/or field/trip blanks  (field  QC) may accompany soil/sediment
          samples and/or water samples that are delivered to a laboratory for
          analysis.  The Contractor shall  not perform MS/MSD analysis on any
          of the field  QC samples.

12.2.2.3  If the EPA Region designates a  sample to be used as an MS/MSD, then
          that sample must be used.  If there is  insufficient sample, less
          than the  required amount, remaining to perform an MS/MSD,  then the
          Contractor shall choose another  sample  to perform an MS/MSD
          analysis.  At the time the selection is made, the Contractor shall
          notify the Region  (through SMO)  that insufficient sample was
          received  and  identify the EPA sample selected for the MS/MSD
          analysis.  The rationale for the choice of a sample other  than the
          one designated by the Region shall be documented in  the  SDG
          Narrative.

12.2.2.4  If there  is  insufficient sample  remaining in any of  the  samples in
          an SDG to perform an MS/MSD, then the Contractor shall  immediately

                                   D-51/VOA                            OLM04.2

-------
Exhibit D Volatiles -- Section 12
Quality Control
Matrix Spike/Matrix Spike Duplicate (MS/MSD)

          contact SMO to inform them of the problem.  SMO will contact the
          Region for instructions.  The Region will either approve that no
          MS/MSD is required, or require that a reduced sample aliquot be used
          for the MS/MSD analysis.  SMO will notify the Contractor of the
          resolution.  The Contractor shall document the decision in the SDG
          Narrative.

12.2.2.5  If it appears that the Region has requested MS/MSD analysis at a
          greater frequency than required by the contract, the Contractor
          shall contact SMO.  SMO will contact the Region to determine which
          samples should have an MS/MSD performed on them.  SMO will notify
          the Contractor of the Region's decision.  The Contractor shall
          document the decision in the SDG Narrative.  If this procedure is
          not followed, the Contractor will not be paid for MS/MSD analysis
          performed at a greater frequency than required by the contract.

12.2.2.6  When a Contractor receives only a Performance Evaluation (PE)
          sample(s), no MS/MSD shall be performed within that SDG.

12.2.2.7  When>a Contractor receives a PE sample as part of a larger SDG, a
          sample other than the PE sample must be chosen for the MS/MSD when
          the Region did not designate samples to be used for this purpose.
          If the PE sample is received as an ampulated standard extract, the
          ampulated PE sample is not considered to be another matrix type.

12 . 2 . 3  Procedure for. Preparing MS/MSD

12.2.3.1  Water

          To prepare a matrix spike and matrix spike duplicate for water
          samples, add 10 uL of the matrix spike solution (Section 7.2.4.2) to
          each of the 5 mL aliquots of the sample chosen for spiking.  Process
          samples according to Sections 10.1.3.7 through 10.1.3.12.
          Disregarding any dilution's, this is equivalent to a concentration of
          50 ug/L of each matrix spike compound.

12.2.3.2  Soil - Low Level

          To prepare a matrix spike and matrix spike duplicate for low level
          soil/sediment samples, add 10 uL of the matrix spike solution to the
          5 mL of spiked reagent water added to each of the two aliquots of
          the soil/sediment from the sample chosen for spiking.  Process
          samples according to Sections 10.1.4.8 through 10.1.4.13.  The
          concentration for a 5 g sample should be equivalent to 50 ug/kg of
          each matrix spike compound.

12.2.3.3  Soil - Medium Level

12.2.3.3.1    To prepare a matrix spike and matrix spike duplicate for medium
              level soil/sediment samples, add 9 mL of methanol and 1 mL of
              matrix spike solution to each of the two aliquots of the
              soil/sediment sample chosen for spiking.  Process samples
              according to Sections 10.1.5.6 through 10.1.5.10.  This results
              in a 6,200 ug/kg concentration of each matrix spike compound
              when added to a 4 g sample.  Add a 100 uL aliquot of this

                                   D-52/VOA                            OLM04.2

-------
                                             Exhibit D Volatiles -- Section 12
                                                              . Quality Control
                                  Matrix Spike/Matrix Spike Duplicate (MS/MSD)

              extract to 5 mL of water for purging (as per Sections 10.1.5.8
              through 10.1.5.9).

12.2.3.3.2    NOTE:  Before performing an MS/MSD analysis, analyze the sample
              used for MS/MSD.  If the sample analysis required dilution,  the
              aliquots for the MS/MSD shall be prepared at the same dilution
              as the least diluted analysis for which the sample results will
              be reported to the Agency.  Sample dilutions must be performed
              in accordance with Section 10.1.6.  Do not further dilute MS/MSD
              samples to get either spiked or non-spiked analytes within
              calibration range.

12.2.4  Calculations for MS/MSD

12.2.4.1  Calculate the concentrations of the matrix spike compounds using the
          same equations as used for target compounds  (Equations 5, 6, and 7).
          Calculate the recovery of each matrix spike compound as follows:
          EQ.  12
                                            SSR ~ SR
                     Matrix Spike Recovery = 	 x 100
                                               SA
          Where,

          SSR = Spiked sample result
          SR  = Sample result
          SA  = Spike added

12.2.4.2  Calculate the relative percent difference  (RPD) of the recoveries of
          each compound in the matrix spike and matrix spike duplicate as
          follows:
          EQ.  13
                           RPD =   IMSR-MSDRI   x 100
                                 — (MSR + MSDR )
                                 2
          Where,

          MSR  = Matrix  spike  recovery
          MSDR = Matrix  spike  duplicate recovery

          The vertical bars  in the  formula above  indicate the absolute value
          of the difference, hence  RPD is always  expressed as a positive
          value.

 12.2.5  Technical Acceptance Criteria for MS/MSD

 12.2.5.1  All MS/MSDs must be  analyzed on a GC/MS system meeting the BFB,
          initial calibration  and continuing calibration technical acceptance
                                  .- D-53/VOA                           OLM04 . 2

-------
Exhibit D Voiatiies -- Section 12
Quality Control
Matrix Spike/Matrix Spike Duplicate  (MS/MSD)

          criteria, the blank technical acceptance criteria, and at the
          frequency described in Section 12.2.2.

12.2.5.2  The MS/MSD must be analyzed within the contract holding time.

12.2.5.3  The retention time shift for each of the internal standards in the
          MS/MSD must be within ±0.50 minutes  (30 seconds) of its retention
          time and the most recent continuing calibration standard analysis.

12.2.5.4  The limits for matrix spike compound recovery and RPD are given in
          Table 8.  As these limits are only advisory, no further action by
          the laboratory is required.  However, frequent failures to meet the
          limits for recovery or RPD warrant investigation by the laboratory,
          and may result in questions- from the Agency.

12.2.6  Corrective Action for MS/MSD

        Any MS/MSD that does not meet the technical acceptance criteria for
        MS/MSD must be reanalyzed at no additional cost to the Agency.
                                   D-54/VOA                            OLM04.2

-------
                                         Exhibit D Volatiles -- Sections 13-16
           Method Performance/Pollution Prevention/Waste Management/References

13.0  METHOD PERFORMANCE

      Not applicable.

14.0  POLLUTION PREVENTION

14.1  Pollution prevention encompasses any technique that reduces or
      eliminates the quantity or toxicity of waste at the point of generation.
      Numerous opportunities for pollution prevention exist in laboratory
      operation.  The EPA has established a preferred hierarchy of
      environmental management techniques that places pollution prevention as
      the management option of first choice.  Whenever feasible, laboratory
      personnel should use pollution prevention techniques to address their
      waste generation.  When wastes cannot be feasibly reduced at the source,
      the Agency recommends recycling as the next best option.

14.2  For information about pollution prevention that may be applicable to
      laboratories and research institutions consult "Less is Better:
      Laboratory Chemical Management for Waste Reduction," available from the
      American Chemical Society's Department of Government Relations and
      Science Policy, 1155 16th Street, N.W., Washington D.C., 20036, (202)
      872-4477.

15.0  WASTE MANAGEMENT

      The Environmental Protection Agency requires that laboratory waste
      management practices be conducted consistent with all applicable rules
      and regulations.  The Agency urges laboratories to protect the air,
      water, and land by minimizing and controlling all releases from hoods
      and bench operations, complying with the letter and spirit of any sewer
      discharge permits and regulations, and by complying with all solid and
      hazardous waste regulations, particularly the hazardous waste
      identification rules and land disposal restrictions.  For further
      information on waste management consult "The Waste Management Manual for
      Laboratory Personnel," available from the American Chemical Society at
      the address listed in Section 14.2.

16.0  REFERENCES

      Not applicable.
                                   .D-55/VOA                            OLM04.2

-------
Exhibit D Volatiles -- Section 17
Tables/Diagrams/Flowcharts

17.0  TABLES/DIAGRAMS/FLOWCHARTS
                                    Table  1
                    BFB Key Ions  and Ion Abundance  Criteria
        Mass Ion Abundance Criteria
            50    8.0-40.0 percent of mass 95
            75    30.0-66.0 percent of mass 95
            95    base peak, 100 percent relative abundance
            96    5.0-9.0 percent of mass 95  (see note)
           173    less than 2.0 percent of mass 174
           174    50.0-120.0 percent of mass 95
           175    4.0-9.0 percent of mass 174
           176    93.0-101.0 percent of mass 174
           177    5.0-9.0 percent of mass 176

        NOTE:  All ion abundances must be normalized to m/z 95, the nominal
        base peak, even though the ion abundance of m/z 174 may be up to 120.0
        percent that of m/z 95.
                                   D-56/VOA                            OLM04.2

-------
                                      Exhibit D Volatiles -- Section 17
                                             Tables/Diagrams/Flowcharts
                             Table 2
        Characteristic Ions for Volatile Target Compounds
Primary
Quantitation
Analvte Ion Secondary Ion(s)
Di chlorodi fl uoromethane
Chloromethane
Vinyl chloride
B romome thane
Chloroethane
Tri chl orofluoromethane
1, 1-Dichloroethene
1, l,2-Trichloro-l,2,2-trifluoroethane
Acetone
Carbon disulfide
Methyl Acetate
Methylene chloride
trans-1, 2-Dichloroethene
Methyl tert -Butyl Ether
1, 1-Dichloroethane
cis-1, 2-Dichloroethene
2-Butanone
Chloroform
1,1, 1-Trichloroethane
Cyclohexane
Carbon Tetrachloride
Benzene
1, 2-Dichloroethane
Trichloroethene
Me thyl cycl ohexane
1, 2-Dichloropropane
Bromodichloromethane
cis-1, 3-Dichloropropene .
4 -Methyl -2 -pentanone
Toluene
trans-1, 3-Dichloropropene
1,1, 2 -Trichloroethane
Tetrachloroethene
2-Hexanone
Dibromochloromethane
1 , 2 -Dibromoethane
Chlorobenzene
Ethylbenzene
Xylene (total)
Styrene
Bromoform

85
50
62
94
64
101
96
101
43
76
43
84
96
73
63
96
43*
83
97
56
117
78
62
130
83
63
83
75
43
91
75
97
164
43
129
107
112
106
106
104
173

87
52
64
96
66
103
61, 98
85, 151
58
78
74
49, 51, 86
61, 98
43, 57
65, 83, 85, 98, 100
61, 98
57
85
99, 117, 119
69, 84
119, 121
-
64, 100, 98
95, 97, 132
55, 98
65, 114
85
77
58, 100
92
77
83, 85, 99, 132, 134
129, 131, 166
58, 57, 100
208, 206
109
114
91
91
78, 103
171, 175, 250, 252, 254,
256
*m/z 43 is used for quantitation of 2-Butanone, but m/z 72 must be
present for positive identification.
                             D-57/VOA
OLM04.2

-------
Exhibit D Volatiles -- Section 17
Tables/Diagrams/Flowcharts
                                Table 2 (Con't)

               Characteristic Ions for Volatile  Target Compounds
      Analvte
   Primary
 Quantitation
	Ion	
Secondary Ion(s)
      Isopropylbenzene
      1,1,2,2-Tetrachloroethane
      1, 3-Dichlorobenzene
      1,4-Dichlorobenzene
      1,2-Dichlorobenzene
      1,2-Dibromo-3-Chioropropane
      1,2,4-Trichlorobenzene
     105
      83
     146
     146
     146
      75
     180
120, 77
85, 131,  133,  166
111, 75
111, 75
111, 75
157, 155
182, 145
                                   D-58/VOA
                                 OLM04.2

-------
                                             Exhibit D Volatiles -- Section 17
                                                    Tables/Diagrams/Flowcharts
                                    Table  3
       Volatile  Internal  Standards  with Corresponding  Target  Compounds
           and System Monitoring Compounds  Assigned  for Quantitation
Bromochloromethane

Dichlorodifluoromethane
Chloromethane           |
Vinyl Chloride
Bromome thane
Chloroethane
Trichlorofluoromethane  \
1,1-Dichloroethene
1,1,2-Trichloro-l, 2,2-
   trifluoroethane
Acetone
Carbon Disulfide
Methyl Acetate
Methylene Chloride
trans-1,2-
   Dichloroethene
Methyl tert-Butyl Ether
1,1-Dichloroethane
cis-l,2-Dichloroethene
2-Butanone
Chloroform
1,2-Dichloroethane
1, 2-D,ichloroethane-d4
    (SMC)
1,4 -Pi fluorobenzene

1,1,1-Trichloroethane
Cyclohexane
Carbon Tetrachloride
Benzene
Trichloroethene         |
Me thylcyclohexane
1,2-Dichloropropane
Bromodichloromethane
cis-1,3-Dichloropropene
trans-1,3-              |
   Dichloropropene
1,1,2-Trichloroethane
Dibromochloromethane
Bromoform
Chlorobenzene-d5

4-Methyl-2-pentanone
Toluene
Tetrachloroethene
2-Hexanone
1,2-Dibromoethane
Chlorobenzene
Ethylbenzene
Xylene  (total)
Styrene
Isopropylbenzene
1,1,2,2-
   Tetrachloroethane
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1,2-Dichlorobenzene
l,2-Dibromo-3-
   ch1oropropane
1,2,4-Trichlorobenzene
Toluene-d8  (SMC)
4-Bromofluorobenzene
    (SMC)
 (SMC)  =  system monitoring compound
                                    D-59/VOA
                                              OLM04.2

-------
Exhibit D Volatiles -- Section 17
Tables/Diagrams/Flowcharts
                                    Table 4
            Characteristic Ions for System Monitoring Compounds and
      Internal Standards for Volatile Organic Compounds with CAS Numbers
Compound
   Primary
 Quantitation
	Ion	
Secondary
  lon(s)
CAS Number
4-Bromofluorobenzene
1,2-Dichloroethane-d4
Toluene-d8
SYSTEM MONITORING COMPOUNDS
     95            174, 176      460-00-4
     65            102           17060-07-0
     98            70, 100       2037-26-5
Bromochloromethane
1,4-Difluorobenzene
Chlorobenzene-d5
     INTERNAL  STANDARDS
     128           49, 130, 51   74-97-5
     114     .      63, 88        540-36-3
     117           82, 119       3114-55-4
                                   D-60/VOA
                                             OLM04.2

-------
                                    Exhibit D Volatiles -- Section 17
                                           Tables/Diagrams/Flowcharts
                          Table  5
Relative Response Factor Criteria for Initial and Continuing
          Calibration of Volatile  Organic  Compounds
Volatile
Compound
Di chlorodi fl uorome thane
Chloromethane
Vinyl chloride
Bromome thane
Chi oroe thane
Tri chl orofl uorome thane
1, 1-Dichloroethene
1, l,2-Trichloro-l,2,2-trifluoroethane
Acetone
Carbon disulfide
Methyl Acetate
Methylene chloride
trans-l,2-Dichloroethene
Methyl tert -Butyl Ether
1, 1-Dichloroethane
cis-l,2-Dichloroethene
2-Butanone
Chloroform
1,1, 1-Trichloroethane
Cyclohexane
Carbon tetrachloride
Benzene
1 , 2 -Dichloroethane
Trichloroethene
Me thyl cycl ohexane
1 , 2 -Dichloropropane
Bromodichloromethane
cis- 1 , 3 -Dichloropropene
4 -Methyl -2 -pentanone
Toluene
trans- 1, 3 -Dichloropropene
1,1, 2 -Trichloroethane
Tetrachloroethene
2-Hexanone
Dibromochloromethane
1, 2-Dibromoe thane
Chlorobenzene
Ethylbenzene
Xylene (total)
Styrene
Bromoform
Minimum
RRF
0.010
0.010
0.100
0.100
0.010
0.010
0.100
0.010
0.010
0.010
0.010
0.010
0.010
0. 010
0.200
0.010
0.010
0.200
0.100
0.010
0.100
0.500
0.100
0.300
0.010
0.010
0.200
0.200
0.010
0.400
0.100
0.100
0.200
0.010
0.100
0.010
0.500
0.100
0.300
0.300
0.100
Maximum
%RSD
none
none
20.5
20.5
none
none
20.5
none
none
none
none
none
none
none
20.5
none
none
20.5
20.5
none
20.5
20.5 .
20.5
20 .5
none
none
20.5
20.5
none
20.5
20 .5
20.5
20.5
none
20.5
none
20.5
20.5
20.5
20.5
20.5
Maximum
%Diff
none
none
±25.0
±25.0
none
none
±25.0
none
none
none
none
none
none
none
±25.0
none
none
±25.0
±25.0
none
±25.0
±25.0
±25.0
±25.0
none
none
±25.0
±25.0
none
±25.0
±25.0
±25.0
±25.0
none
±25.0
none
±25.0
±25.0
±25.0
±25.0
±25.0
                           D-61/VOA
OLM04.2

-------
Exhibit D Volatiles -- Section 17
Tables/Diagrams/Flowcharts

                                Talile 5 (Con't)

         Relative Response Factor Criteria for Initial and Continuing
                   Calibration of Volatile Organic  Compounds
        Volatile
        Compound
Minimum  Maximum  Maximum
RRF	%RSD     %Diff
        Isopropylbenzene
        1,1,2,2-Tetrachloroethane
        1,3-Dichlorobenzene
        1,4-Dichlorobenzene
        1,2-Dichlorobenzene
        1, 2-Dibromo-3-chloropropane
        1,2,4-Trichlorobenzene

        SYSTEM MONITORING COMPOUNDS
        Bromofluorobenzene
        Toluene-d8
        1,2-Dichloroethane-d4
0.010
0.300
0.600
0.500
0.400
0. 010
0.200
none
20.5
20.5
20.5
20.5
none
20.5
none
±25.0
±25.0
±25. 0
±25.0
none
±25.0
0.200
0.010
0.010
20.5
none
none
±25.0
none
none
                                   D-62/VOA
                       OLM04.2

-------
                                     Exhibit D Volatiles  --  Section  17
                                            Tables/Diagrams/Flowcharts
                            Table 6

                      The  "X"  Factor  Table
X Factor
Estimated
Concentration Range1
(ug/kg)
Take This Volume of
Methanol Extract2 (uL)
0.25
0.5 -
2.5 -
12.5
- 5.0
10.0
50.0
- 250
500 - 10,000
1000 - 20,000
5000 - 100,000
25,000 - 500,000
100
50
10
100 of



1/50 dilution3
Calculate appropriate dilution factor for concentrations exceeding
those in the table.
Actual  concentration ranges  could be  10  to 20  times  higher than this
if  the  compounds  are halogenated and  the estimates are  from GC/FID.

The volume  of  methanol  added to the 5 mL of water being purged should
be  kept constant.  Therefore,  add to the  5 mL syringe whatever volume
of  methanol is necessary to  maintain  a volume  of  100 uL added to the
syringe.

Dilute  an aliquot of the methanol extract and  then take 100 uL for
analysis.
                            D-63/VGA
                                                OLM04.2

-------
Exhibit D Volatiles -- Section 17
Tables/Diagrams/Flowcharts
                                    Table  7

                  System Monitoring Compound Recovery Limits
             Compound                % Recovery Water    % Recovery Soil

             Toluene-d8                   88-110              84-138
             Bromofluorobenzene           86-115              59-113
             l,2-Dichloroethane-d4        76-114              70-121
                                    Table  8

                           Matrix Spike  Recovery and
                      Relative Percent Difference Limits
                              % Recovery   RPD      % Recovery  RPD
        Compound              Water        Water    Soil        Soil

        1,1-Dichloroethene    61-145       14       59-172      22
        Trichloroethene       71-120       14       62-137      24
        Benzene               76-127       11       66-142      21
        Toluene               76-125       13       59-139      21
        Chlorobenzene         75-130       13       60-133      21
                                   D-64/VOA                            OLM04.2

-------
                                             Exhibit D -- Volatiles Appendix A
                                Screening of Hexadecane Extracts for Volatiles

APPENDIX A - SCREENING OF HEXADECANE EXTRACTS FOR VOLATILES

1.0   SCOPE AND APPLICATION

1.1   The hexadecane extraction and screening methods for purgeables described '
      in this section are designed to aid the analyst in deciding whether.a
      soil sample is low or medium level in order to prevent saturation of the
      purge and trap system and/or the GC/MS system.  These or other screening
      methods should be used, particularly if there is some doubt about the
      level of organics in a sample.  This is especially true in soil/sediment
      analysis.  Water samples may also be screened to determine an
      appropriate dilution factor for analysis.

1.2   These extractions and preparation procedures were developed for rapid
      screening of water and soil/sediment samples from hazardous waste sites.
      The design of the methods thus does not stress efficient recoveries or
      low limits of quantitation.  Rather, the procedures were designed to
      screen at moderate recovery and sufficient sensitivity for a broad
      spectrum of organic chemicals.  The results of the analyses thus may
      reflect only a minimum of the amount actually present in some samples.
      This is especially true if water soluble solvents are present.

2.0   SUMMARY OF METHODS

2.1   Sample Preparation

2.1.1    Water

         A 40 mL aliquot of  sample  is extracted with 2 mL of hexadecane.  This
         provides a minimum  quantitation  limit  (MQL) as follows:

            Compounds                           MQL  (uq/L)
            non-halogenated  aromatics           40-50
            halogenated methanes     -          80-1000
            halogenated ethanes                 400-500

2.1.2    Soil/Sediment

         40 mL of reagent water are added to  10 g  (wet weight) of
         soil/sediment and shaken.  The water phase is in turn extracted  with
         2 mL of hexadecane.  This  provides a minimum quantitation  limit  of
         approximately four  times higher  than those listed  for water.

2.2   GC/FID Screening

      The hexadecane  extracts of water and  soil/sediment are screened on  a  gas
      chromatograph/flame  ionization detector (GC/FID).  The results of the
      screen will determine  if volatile organics are to be  analyzed by low  or
      medium level GC/MS procedures if the  sample  is a soil/sediment, or  to
      determine  the appropriate dilution  factor if  the sample is water.   Note:
      The flame  ionization detector varies  considerably  in  sensitivity when
      comparing  aromatics  and halogenated methanes  and ethanes.  Halomethanes
      are approximately  2Ox  less sensitive  than aromatics and haloethanes are
                                    D-65/VOA                            OLM04.2

-------
Exhibit D -- Volatiles Appendix A
Screening of Hexadecane Extracts for Volatiles

      approximately lOx less sensitive than aromatics.  Low molecular weight,
      water soluble solvents (e.g., alcohols and ketones) will not be
      extracted from the water, and therefore will not be detected by the
      GC/FID.

3.0   INTERFERENCES

      Method interferences may be caused by contaminants in solvents,
      reagents, glassware, and other sample processing hardware that lead to
      discrete artifacts and/or elevated baselines in the total ion current
      profiles.  All of these materials must be routinely demonstrated to be
      free from interferences under the conditions of the analysis by running
      laboratory reagent blanks.  Matrix interferences may be caused by
      contaminants that are coextracted from the sample.  The extent of matrix
      interferences will vary considerably from source to source depending
      upon the nature and diversity of the site being sampled.

4.0   SAFETY

      The toxicity or carcinogenicity of each reagent used in this method has
      not been precisely determined; however, each chemical should be treated
      as a potential health hazard.  Exposure to these reagents should be
      reduced to the lowest possible level.  The laboratory is responsible for
      maintaining a current awareness file of OSHA regulations regarding the
      safe handling of the chemicals specified in this method.  A reference
      file of data handling sheets should be made available to all personnel
      involved in these analyses.  Use all reagents in fume hoods whenever
      possible.  Always wear safety glasses or a shield for eye protection,
      protective clothing, and observe proper mixing when working with these
      reagents.

5.0   EQUIPMENT AND SUPPLIES

      Brand names, suppliers, catalog and part numbers are for illustrative
      purposes only.  No endorsement is implied.  Equivalent performance may
      be achieved using equipment and supplies other than those specified
      here, but demonstration of equivalent performance meeting the
      requirements of this SOW is the responsibility of the Contractor.

5.1   Glassware

5.1.1    Syringes - 0.5 mL

5.1.2    Vials and Caps - 2 mL capacity for GC autosampler

5.1.3    Pasteur Pipets - disposable

5.1.4    Centrifuge Tube - 50 mL with ground glass stopper or Teflon-lined
         screw cap.

5.1.5    Volumetric Flask - 50 mL with ground glass stopper.

5.2   Balance  - analytical, capable of accurately weighing ±0.0001 g.

5.3   Pyrex Glass Wool
                                   D-66/VOA                            OLM04.2

-------
                                             Exhibit D -- Volatiles Appendix A
                                Screening of Hexadecane Extracts for Volatiles

5.4   Balances - analytical, capable of accurately weighing ±0.0001 g,  and a
      top-loading balance capable of weighing 100 g ±0.01 g.   The balances
      must be calibrated with class S weights or known reference weights once
      per each 12-hour work shift.  The balances must be calibrated with class
      S weights at a minimum of once per month.  The balances must also be
      annually checked by a certified technician.

5.5   Centrifuge  •

5.6   Gas Chromatograph/Mass Spectrometer (GC/MS) System

5.6.1    Gas Chromatograph  - an analytical system complete with a temperature
         programmable gas chromatograph suitable for on-column injection and
         all required accessories including syringes, analytical columns,
         gases, detector, and strip-chart recorder.  A data system is
         recommended for measuring peak areas.

5.6.2    Gas Chromatography Column - 30 m (or longer) x 2 mm ID glass column
         packed with 10% OV-101 on 100-120 mesh chromosorb W-HP  (or
         equivalent).  The  column temperature should be programmed from 80 °C
         to 280 °C at 16 C°/min and held at  280 °C for 10 minutes.

5.6.3    Flame lonization Detector

6.0   REAGENTS AND STANDARDS

6.1   Reagents

6.1.1    Reagent Water - defined  as water in which'an interferant is not
         observed at the CRQL of  each  analyte of  interest.

6.1.2    Hexadecane  and Methanol  - pesticide  residue analysis grade or
         equivalent.

6.2   Standards

6.2.1    Introduction

         The  Contractor must provide all standard solutions  to be used with
         this  contract.  These  standards may  be  used only after  they have been
         certified according to the procedure  in Exhibit  E.  The Contractor
         must  be able to verify that the standards  are  certified.
         Manufacturer's certificates of analysis  must be  retained by the
         contractor  and presented upon request.

6.2.2    Stock Standard Solutions

6.2.2.1      Stock standard  solutions  (1 ug/uL)  can be prepared  from pure
             standard materials  or purchased as  certified solutions.
                                    D-67/VOA                            OLM04.2

-------
Exhibit D -- Volatiles Appendix A
Screening of Hexadecane Extracts for Volatiles

6.2.2.2     Prepare stock standard solutions by accurately weighing about 0.01
            g of pure material.  Dissolve the material in methanol and dilute
            to volume in a 10 mL volumetric flask.  Larger volumes can be used
            at the convenience of the analyst.  If compound purity is
            certified at 97% or greater, the weight can be used without
            correction to calculate the concentration of the stock standard.

6.2.3    Working Standard Solutions

6.2.3.1     Standard Mixture #1

            Prepare a working standard mixture containing benzene, toluene,
            ethylbenzene, and xylene at 100 ng/uL of each compound in
            methanol.

6.2.3.2     Standard Mixture #2

            Prepare a working standard mixture containing n-nonane.  And n-
            dodecane at 100 ng/uL of each compound in methanol.

6.2.4    Storage of Standards

         Transfer all standard solutions into multiple Teflon-sealed screw-cap
         vials.  Store, with no head-apace, at -10 °C to -20 °C,  and protect
         from light.  Stock standard solutions should be.checked frequently
         for signs of degradation or evaporation, especially just prior to
         preparing calibration standards from them.  These solutions must be
         replaced after six months, or sooner, if comparison with quality
         control check samples indicates a problem.  Standards prepared from
         gasses or reactive compounds such as styrene must be replaced after
         two months, or sooner, if comparison with quality control check
         samples indicates a problem.

7.0   QUALITY CONTROL

7.1   Method Blank

7.1.1    Summary

         A method blank is a volume of clean reagent water taken through the
         extraction and screening procedure.  The volume of reagent water -used
         must be approximately equal to the volume of associated samples.  The
         purpose of a method blank is to determine the levels of contamination
         associated with the processing and analysis of samples.

7.1.2    Frequency

         One method blank must be extracted and analyzed on each GC/FID system
         used to screen samples for the following, whichever is most frequent.

         •  Each SDG, or
                                   D-68/VOA                            OLM04.2

-------
                                             Exhibit D -- Volatiles Appendix A
                                Screening of Hexadecane Extracts for Volatiles

         •  Each 20 samples.in an SDG,  excluding matrix spikes/matrix spike
            duplicates,  or

         •  When samples are extracted.

7.1.3    Procedure

         For screening of volatile organics, a method blank consists of a 40
         mL volume of reagent water extracted with 2 mL of hexadecane.   The
         hexadecane extract is then screened on a GC/FID system.

8.0   CALIBRATION AND STANDARDIZATION

8.1   GC/FID Operating Conditions

      Refer to Section 5.5.2 for recommended column temperature program.

8.2   GC Calibration

8.2.1    Summary

         Prior to sample analysis each GC/FID system must be standardized for
         half scale response.

8.2.2    Frequency

         Each GC/FID system must be calibrated at the beginning of each 12-
         hour shift.

8.2.3    Procedure

8.2.3.1     Add 200 uL of each of working standard mixtures #1 and #2  (Section
            6.2.3) to separate 40 mL portions of reagent water in 50 mL
            volumetric flasks.   Immediately add 2 mL of hexadecane, cap the
            flask, and shake vigorously  for 1 minute.  Let phases separate.
            Open the flask  and add sufficient reagent water to bring the
            hexadecane layer into the  neck  of the flask.  Transfer
            approximately 1 mL of the  hexadecane layer to a 2 mL GC vial.

8.2.3.2     Inject 1-2 uL of the extracts that contain approximately 10 ng/uL
            each of standard mixture #1  and standard mixture #2 compounds.

9.0    PROCEDURE

9.1    Sample Preparation

9.1.1    Water

9.1.1.1     Allow.the contents of the  40 mL sample  vial  to come to  room
            temperature.'  Quickly transfer  the contents  of the  40 mL sample
            vial to a 50 mL volumetric flask.  Immediately add  2 mL of
            hexadecane, cap the  flask, and  shake vigorously for 1 minute.  Let
            phases separate.  Open the flask  and add  sufficient reagent water
            to  bring the hexadecane  layer into the  neck  of the  flask.
                                    D-69/VOA                            OLM04.2

-------
Exhibit D -- Volatiies Appendix-A
Screening of Hexadecane Extracts for Volatiies

9.1.1.2     Transfer approximately 1 mL of the hexadecane layer to a 2 mL GC
            vial.  If an emulsion is present after shaking the sample, break
            it by doing the following:

            •     Pulling the emulsion through a small plug of Pyrex glass
                  wool packed in a pipet, or

            •     Transferring the emulsion to a centrifuge tube and
                  centrifuging for several minutes.

9.1.2    Soil/Sediment

         Add approximately 10 g.of soil/sediment  (wet weight) to 40 mL of
         reagent water in a 50 mL centrifuge tube with a ground glass stopper
         or Teflon-lined cap.  Cap and shake vigorously for 1 minute.
         Centrifuge the capped flask briefly.  Quickly transfer supernatant
         water to a 50 mL volumetric flask equipped with a ground-glass
         stopper.  Follow Section 9.1.1 starting with the addition of 2 mL of
         hexadecane.

9.2   GC/FID Analysis

      Inject the same volume of sample hexadecane extract as the extracted
      standard mixture in Section 8.2.3.

9.2.1    GC/FID Chromatogram Interpretation -- Following are two options for
         interpreting the GC/FID Chromatograms.

9.2.1.1     Option A is to use standard mixture #1 containing the aromatics to
            calculate an approximate concentration of the aromatics in the
            sample.  Use this information to determine the proper dilution for
            purge and trap if the sample is water, or whether to use the low
            or medium level GC/MS purge and trap methods if the sample is
            soil/sediment  (see Table 1, Section 9.3 for guidance).  This
            should be the best approach; however, the aromatics may be absent
            or obscured by higher concentrations of other purgeables.  In
            these cases, Option B may be the best approach.

9.2.1.2     Option B is to use standard mixture #2 containing n-nonane and n-
            dodecane to calculate a factor.  Use the factor.to calculate a
            dilution for purge and trap of a water sample or to determine
            whether to use the low or medium level GC/MS purge and trap
            methods for soil/sediment samples  (see Table 1, Section 9.3 for
            guidance).  All purgeables of interest have retention times less
            than the n-dodecane.

9.3   Analytical Decision Point

9.3.1    Water

9.3.1.1     Compare the chromatograms of the hexadecane extract of the sample
            with those of the reagent blank and extract of the standard.

9.3.1.2     If no peaks are noted, other than those also in the reagent blank,
            analyze a 5 mL water sample by purge and trap GC/MS.
                                   D-70/VOA                           " OLM04.2

-------
                                             Exhibit D -- Volatiles Appendix A
                                Screening of Hexadecane Extracts for Volatiles

9.3.1.3     If peaks are present prior to the n-dodecane and the aromatics are
            distinguishable, follow Option A (Section 9.2.1).

9.3.1.4     If peaks are present prior to the n-dodecane but the aromatics are
            absent or indistinguishable, use option B as follows:  if all
            peaks are <3% of the n-nonane, analyze a 5 mL water sample by
            purge and trap GC/MS.   If any peaks are 23%of the n-nonane,
            measure the peak height or area of the major peak and calculate
            the dilution factor as follows:
               .,   ,   _        Peak area of sample major peak
              Dilution Factor =	 x 50
                                   Peak area of n-nonane
            The water sample will be diluted using the calculated factor just
            prior to purge and trap GC/MS analysis.

9.3.2    Soil/Sediment

9.3.2.1     Compare the chromatograms of the hexadecane extract of the sample
            with those of the reagent blank and extract of the standard.

9.3.2.2     If no peaks are noted, other than those also in the reagent blank,
            analyze a 5 g sample by low level GC/MS.

9.3.2.3     If peaks are present prior to the n-dodecane and the aromatics are
            distinguishable, follow Option A  (Section 9.2.1) and the
            concentration information in Table 1, to determine whether to
            analyze by low or medium level method.

9.3.2.4     If peaks are present prior to the n-nonane but the aromatics are
            absent or indistinguishable, and using Option B as follows,
            calculate a factor using the following formula:

                            _ Peak area of sample major peak
                                  Peak area of n-nonane
                                    Table 1
                  Determination of  GC/MS  Purge  and Trap Method
                                                Approximate  Concentration
             X  Factor    Analyze  by             Range*  (ug/Kg)

             0-1.0        low level method       0-1,200
             >  1.0        medium level method    >1,200
          *   This  concentration range is based on the  response of  aromatics  to
          GC/FID.   When comparing GC/FID responses,  the  concentration for
          halomethanes is 20 times higher,  and that  for  haloethanes is 10  times
          higher.
                                    D-71/VOA                            OLM04.2

-------
Exhibit D -- Volatiles Appendix 3
Modified SW-846 Method 5035 for Volatiles in Low Level Soils

APPENDIX B - MODIFIED SW-846 METHOD 5035 FOR VOLATILES IN LOW LEVEL SOILS

1.0   SCOPE AND APPLICATION

1.1   The analytical method that follows is designed to analyze low level
      sediment and soil samples from hazardous waste sites for the volatile
      organic compounds on the Target Compound List (TCL, see Exhibit C) .  The
      method includes sample preparation, screening to determine the
      approximate concentration of organic constituents in the sample, and the
      actual analysis which is based on a closed-system purge and trap gas
      chromatograph/mass spectrometer (GC/MS) method.

2.0   SUMMARY OF METHOD

2.1   Low level volatile organic compounds are determined by analyzing
      approximately 5 g of sample, in a pre-weighed vial with a septum-sealed
      screw-cap  (see Section 5.0) that already contains a stirring bar and a
      sodium bisulfate preservative solution. Note: The sodium bisulfate
      preservative and the stirring bar may be omitted under certain
      circumstances  (see Sections 9.3.2 and 9.3.8).  The entire vial is placed
      into the instrument carousel.  Immediately before analysis, organic-free
      reagent water, surrogates, and internal standards are 'automatically
      added without opening the sample vial.  The vial containing the sample
      is heated to 40 "C and the volatiles purged through a sorbent trap using
      an inert gas combined with agitation of the sample.  When purging is
      complete, the' trap is heated and backflushed with helium to desorb the
      purgeable compounds onto a gas chromatograph column.  The gas
      chromatograph is temperature-programmed to separate the purgeable
      compounds which are then detected with a mass spectrometer.

2.2   The sample introduction technique in Section 2.1 is not applicable to
      all samples.  If sample screening indicates that the soil/sediment
      sample should be analyzed as a medium level .sample, the Contractor shall
      follow the procedure described in Exhibit D-VOA Section 10.1.5 for
      medium level soil/sediment samples.

3. 0   INTERFERENCES

3.1   Method interference may be caused by impurities in the purge gas,
      organic compounds out-gassing from the plumbing ahead of the trap, and
      solvent vapors in the laboratory.   The analytical system must be
      demonstrated to be free from contamination under the conditions of the
      analysis by running laboratory method and instrument blanks as described
      in Exhibit D-VOA Section 12.  The use of non-Polytetrafluoroethylene
       (PTFE) tubing, non-PTFE thread sealants, or flow controllers with rubber
      components in the purging device should be avoided.

3.2   Samples can be contaminated by diffusion of volatile organics
       (particularly fluorocarbons and methylene chloride) through the septum
      seal into the sample during storage and handling.

3.3   Contamination by carryover can occur whenever medium level and low level
      samples are sequentially analyzed.  For samples containing large amounts
      of water-soluble materials, suspended solids, high-boiling compounds, or
      high purgeable levels, it may be necessary to wash out the purging
      device.  The trap and other parts of the system are also subjected to

                                  •: D-72/VOA                            OLM04 .2

-------
                                             Exhibit D -- Volatiles Appendix B
                  Modified SW-846 Method 5035 for Volatiles in Low Level Soils

      contamination; therefore,  frequent bakeout and purging of the entire
      system may be required.

3.4   The laboratory where volatile analysis is performed should be completely
      free of solvents.  Special precautions must be taken, to determine
      methylene chloride.  The analytical and sample storage area should be
      isolated from all atmospheric sources of methylene chloride,  otherwise
      random background levels will result.  Since methylene chloride will
      permeate through PTFE tubing, all GC carrier gas lines and purge gas
      plumbing should be constructed of stainless steel or copper tubing.
      Laboratory workers' clothing previously exposed to methylene chloride
      fumes during common liquid/liquid extraction procedures can contribute
      to sample contamination.  The presence of other organic solvents in the
      laboratory where volatile organics are analyzed will also lead to random
      background levels and the same precautions must be taken.

4 . 0   SAFETY

4.1   The toxicity or carcinogenicity of each reagent used in this method has
      not been precisely defined; however, each chemical compound should be
      treated as a potential health hazard.  From this viewpoint, exposure to
      these chemicals must be reduced to the lowest possible level by whatever
      means available.  The laboratory is responsible for maintaining a
      current awareness file of OSHA regulations regarding the safe handling
      of the chemicals specified in this method.  A reference file of material
      data handling sheets should also be made available to all personnel
      involved in the  chemical analysis.

4.2   The following analytes covered by this method have been tentatively
      classified as known or suspected, human or mammalian carcinogens:
      benzene, carbon  tetrachloride, chloroform, and vinyl chloride.  Primary
      standards of  these toxic compounds should be prepared in a hood.  A
      NIOSH/Mass approved toxic gas respirator should be worn when the analyst
      handles high  concentrations of these toxic compounds.

5.0   EQUIPMENT AND SUPPLIES

      Brand .names,  suppliers, and part numbers are  for illustrative purposes
      only.  No endorsement  is implied.  Equivalent performance may be
      achieved using  equipment and supplies  other than those specified here,
      but demonstration  of equivalent performance meeting  the requirements of
      this SOW is the responsibility of  the  Contractor.  The Contractor  shall
      document any  use of alternate equipment or supplies  in the SDG
      Narrative.

5.1   Sample Containers

      The specific  sample containers required will  depend  on the purge-and-
      trap system  to  be  employed.  Several systems  are commercially available.
      Some systems  employ 40-mL  clear  vials  with a  special  frit  and equipped
      with  two PTFE-faced silicone septa.  Other systems permit  the use  of any
      good quality glass vial  that is  large  enough  to  contain  at least  5 g of
      soil or solid material and at least  10 mL of  water,  and  that can be
      sealed with  a screw-cap containing a PTFE-faced  silicone  septum.   The
      Contractor shall consult  the purge-and-trap system manufacturer's
                                    D-73/VOA                           OLM04.2

-------
Exhibit D -- Volatiles Appendix B
Modified SW-846 Method 5035 for Volatiles in Low Level Soils

      instructions regarding the suitable specific vials, septa, caps, and
      mechanical agitation devices.

5.2   Glassware

5.2.1    Syringes - 25 mL glass hypodermic syringes with Luer-Lok  (or
         equivalent) tip  (other sizes are acceptable depending on sample
         volume used).  5.0, 1.0, and 0.5 mL syringes gas-tight with shut-off
         valve.

5.2.2    Syringe valve - 2-way with Luer ends.

5.2.3    Micro syringes - 25 /uL with a 2 inch x 0.006 inch ID,' 22° bevel
         needle  (Hamilton '#702N or equivalent) .  10 and 100 x/L-

5.2.4    60-mL, septum-sealed glass vials to collect samples for screening,
         dry weight determination.

5.2.5    40-mL, screw-cap, PTFE lined, septum-sealed glass vials.  Examine
         each vial prior to use to ensure that the vial has a flat, uniform
         sealing surface.

5.2:6    Volumetric flasks - Class A, 10-mL and 100-mL, with ground glass
         stoppers.

5.2.7    Disposable Pasteur pipettes.

5.3   Magnetic stirring bars - PTFE- or glass-coated, of the appropriate size
      to fit the sample vials.  Consult manufacturer's recommendation for
      specific stirring bars.  Stirring bars may be reused, provided that they
      are thoroughly cleaned between uses.  Consult the manufacturers of the
      purging device and the stirring bars for suggested cleaning procedures.

5.4   Balances - analytical, capable of accurately weighing ± 0.0001 g, and a
      top-loading balance capable of weighing 100 g ± 0.01 g.  The balances
      must be calibrated with class S weights or known reference weights once
      per each 12-hour work shift.  The balances must be calibrated with class
      S weights at a minimum of once per month.  The balances must also be
      annually checked by a certified technician.

5.5   Purge and Trap Device - consists of a unit that automatically adds
      water, SMCs, and internal standards to a hermetically sealed vial
      containing the sample, purges the volatile compounds using an inert gas
      stream while agitating the contents of the vial, and also traps the
      released volatile compounds for subsequent desorption into the gas
      chromatograph.  Such systems are commercially available from several
      sources and shall meet the following specifications.

5.5.1    The purging device should be capable of accepting a vial sufficiently
         large to contain a 5 g soil/sediment sample plus a magnetic stirring
         bar and 10 mL of water.  The device must be capable of heating a soil
         vial to 40°C and holding it at that temperature while the inert purge
         gas is allowed to pass through the sample.  The device should also be
         capable of introducing at least 5 mL of organic-free reagent water
         into the sample vial while trapping the displaced headspace vapors.
         It must also be capable of agitating the sealed sample during

                                   D-74/VOA                            OLM04.2

-------
                                             Exhibit D -- Volatiles Appendix B
                  Modified SN-846 Method 5035 for Volatiles in Low Level Soils

         purging,  (e.g., using a magnetic stirring bar, sonication, or other
         means).   The analytes being purged must be quantitatively transferred
         to an absorber trap.  The trap must be capable of transferring the
         absorbed volatile compounds to the gas chromatograph.

5.5.2    The trap must be at least 25 cm long and have an inside diameter of
         at least 0.105 inch.  Starting from the inlet, the trap must contain
         equal amounts of the absorbents listed below.  It is recommended that
         1.0 cm of methyl silicone-coated packing  (35/60) mesh  (Davison, grade
         15 or equivalent) be inserted at the inlet to extend the life of the
         trap.

         •  2,6-Diphenylene  oxide polymer - 60/80 mesh, chromatographic grade
            (Tenax GC or equivalent)

         •  Methyl silicone  packing - OV-1  (3%) on Chromosorb-W, 60/80 mesh or
            equivalent

         •  Coconut charcoal - Prepare from Barnebey  Cheney, CA-580-26, or
            equivalent, by crushing through 26 mesh screen

         Trapping materials  other than those listed above may also be used,
         provided that  they  meet the specifications listed in Exhibit D-VOA
         Sections 6.4.2 and  6.4.4.

5.5.3    The desorber for the trap must be  capable of rapidly heating the trap
         to 180  "C.  The polymer section of the trap  should not be heated
         higher than 180  °C  and the remaining sections should not exceed 220
          "C during bakeout mode.

5.6   Gas  Chromatograph/Mass Spectrometer  (GC/MS)  System

5.6.1    Gas chromatograph/mass spectrometer system specifications and
         requirements are described in Exhibit D-VOA  Section  6.6.

6.0   REAGENTS AND STANDARDS

6.1   Reagents

6.1.1    Reagent  water  -  defined as water  in which an interferant  is not
         observed at or above  the CRQL of .the  analytes of interest.  Reagent
         water may be generated by passing tap water  through  a'carbon filter
         bed  containing about  453 g  (1 Ib)  of  activated carbon  (Calgon  Corp.,
         Filtrasorb-300 or  equivalent; .

6.1.1.1     A  water purification system  (Millipore Super-Q or equivalent) may
            be used  to  generate reagent  water.

6.1.1.2     Reagent  water may also be prepared by boiling water for 15
            minutes.   Subsequently,  while  maintaining the temperature at  90
             °C, bubble  a  contaminant-free  inert gas through the water for one
            hour.   While still  hot,  transfer the  water to a narrow-mouth
            screw-cap  bottle and  seal  with a  Teflon-lined septum and cap.

6.1.2    Methanol - pesticide quality  or equivalent.


                                    D-75/VOA                            OLM04.2

-------
Exhibit D -- Volatiles Appendix B
Modified SW-846 Method 5035 for Volatiles in Low Level Soils

6.1.3    Sodium bisulfate - ACS reagent grade or equivalent.

6.2   Standards

6.2.1    The Contractor must provide all standards to be used with this
         contract.  These standards may be used only after  they have been
         certified according to the procedure in Exhibit E.  The Contractor
         must be able to verify that the standards are certified.
         Manufacturer's certificates of analysis must be retained by the
         Contractor and presented upon request.

6.2.2    The Contractor shall follow the procedures described in Exhibit D-VOA
         Section 7.2 for preparing stock standards, secondary dilutions, and
         all working standard solutions.

6.3   Storage of Standard Solutions

6.3.1    The Contractor shall follow the procedures described in Exhibit D-VOA
         Section 7.3 for storage of all standard solutions.

6.3.2    The Contractor is responsible for maintaining the  integrity of
         standard solutions and verifying prior to use.  This means that
         standards must be brought to room temperature prior to use, checked
         for losses, and checked that all components have remained in the
         solution.

7.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE

7.1   Sample Collection and Preservation

7.1.1    Soil/sediment samples should be collected in field core
         sampling/storage containers  (i.e., EnCore™ or equivalent) and  60  mL
         septum-sealed glass vials in sufficient quantity to perform the
         analysis.  The field core sampling/storage containers should contain
         approximately 5 g of sample each.  The Contractor  shall transfer  the
         contents of the field core sampling container immediately upon
         receipt into the closed-system sample vial prepared as described  in
         Section 9.3 below and record the date and time of  transfer.  The
         specific requirements for site sample collection are outlined by  the
         Region.  If soil/sediment samples are received in  pre-prepared
         closed-system purge-and-trap sample vials as described in Section
         9.3, then the Contractor shall proceed to Section  9.3.9 and determine
         final sample weight.

7.1.2    All samples must be iced or refrigerated at 4  °C  (±2 °C) from the
         time of collection until analysis.

7.2   Procedure for Sample Storage

7.2.1    The samples must be protected from light and refrigerated at 4  °C
          (±2  °C) from the time of receipt until 60 days after delivery of  a
         reconciled, complete sample data package to the Agency.  After 60
         days,, the samples may be disposed of in a manner that complies with
         all applicable regulations.
                                   D-76/VOA                            OLM04.2

-------
                                             Exhibit D -- Volatiles Appendix B
                  Modified SW-846 Method 5035 for Volatiles in Low Level  Soils

7.2.2    The samples must be stored in an atmosphere demonstrated to be free
         of all potential contaminants and in a refrigerator used only for
         storage of volatile samples.

7.2.3    All volatile samples in an SDG must be stored together in the same
         refrigerator.

7.2.4    Storage blanks shall be stored with samples until all samples are
         analyzed.

7.2.5    Samples, sample extracts, and standards must be stored separately.

7.2.6    Volatile standards must be stored separately from semivolatile and
         pesticide/Aroclor standards.

7.3   Contract Required Holding Times

      Analysis of soil/sediment samples must be completed within 10 days of
      Validated Time of Sample Receipt  (VTSR).  As part of the Agency's QA
      program, the Agency may provide Performance Evaluation  (PE) samples as
      standard extracts which the Contractor is required to prepare per the
      instructions provided by the Agency.  PE samples must be prepared and
      analyzed concurrently with  the samples in the SDG.  The contract
      required 10 day holding time does not apply to PE samples received as
      standard extracts.

8.0   CALIBRATION AND STANDARDIZATION

8.1   Purge and  Trap

8.1.1    Assemble a purge-and-trap device that meets the specification in
         Section 5.5 and that is  connected to a gas chromatograph/mass
         spectrometer system.

8.1.2    Before  initial use, condition  the trap overnight at  180°C by
         backflushing with an inert gas flow of at least 20 mL/min, or
         according  to the manufacturer's recommendations.  Vent  the trap
         effluent to  the hood, not to the analytical column.  Prior to daily
         use,  condition the trap  for 10 min at 180°C while backflushing.   The
         trap  may be  vented to the analytical column during daily
         conditioning; however, the column must be run  through the•temperature
         program prior to analysis of samples.

8.1.3    Establish  the purge-and-trap instrument operating conditions.  Adjust
         the instrument to inject 5 mL  of reagent water, to heat the sample to
         40°C, and  to hold the sample at 40°C for 1.5 minutes before
         commencing the purge process,  or as recommended by the  instrument
         manufacturer.  Once established, the same purge and  trap conditions
         must  be used for the analysis  of all standards, samples, and blanks.

8.2   Gas  Chromatograph/Mass Spectrometer

      The  Contractor shall follow the instrument conditions described in
      Exhibit  D-VOA Sections 9.1.2 and  9.1.3.
                                   D-77/VOA                            OLM04.2

-------
Exhibit D -- Volatiles Appendix B
Modified SW-846 Method 5035 for Volatiles in Low Level Soils

8.3   GC/MS Calibration  (Tuning) and Ion Abundance

      The Contractor shall follow the procedure described in Exhibit D-VOA
      Section 9.2.  All  technical acceptance criteria for the GC/MS
      performance check  shall be met before any standards or samples,
      including MS/MSDs  or required blanks, are analyzed.  Any samples or
      required blanks analyzed when tuning technical acceptance criteria have
      not been met will  require reanalysis at no additional cost to the
      Agency.

8.4   Initial Calibration

      The Contractor shall follow the procedure described in Exhibit D-VOA
      Section 9.3.  However, the volume of reagent water used for calibration
      must be the same volume used for sample analysis  (normally 5 mL added to
      the vial before sample addition plus the reagent water added by the
      instrument).  The  calibration standards should also contain
      approximately the  same amount of the sodium bisulfate preservative as
      the sample  (e.g.,  approximately 1 g), as the presence of the
      preservative will  affect the purging efficiencies of the analytes.  The
      internal standard  solution must be added automatically, by the
      instrument, in the same fashion as used for the samples.  Place the soil
      vial containing the solution in the instrument carousel.  In order to
      calibrate the system monitoring compounds  (SMCs) using standards at five
      concentrations, it may be necessary to disable the automatic addition of
      SMCs to each vial  containing a calibration standard  (consult the
      manufacturer's instructions).  Prior to purging, heat the sample vial to
      40 "C for 1.5 minutes, or as recommended by the manufacturer.

      All technical acceptance criteria for GC/MS initial calibration
      specified in Exhibit D-VOA Section 9.3.5 shall be met prior to the
      analysis of any samples, including MS/MSDs or required blanks.  Any
      samples or required blanks analyzed when initial calibration technical
      acceptance criteria have not been met will require reanalysis at no
      additional cost to the Agency.

8.5   Continuing Calibration

      The Contractor shall follow the procedure for continuing calibration
      described in Exhibit D-VOA Section 9.4.  However, the continuing
      calibration standard shall be prepared in the same manner as the initial
      calibration standard of the same concentration as specified in Section
      8.4 above  (i.e., addition of the sodium bisulfate preservative).

      All technical acceptance criteria for continuing calibration specified
      in Exhibit D-VOA Section 9.4.5 shall be met prior to the analysis of any
      samples, including MS/MSDs or required blanks.  Any samples or required
      blanks analyzed when continuing calibration technical acceptance
      criteria have not  been met will require reanalysis at no additional cost
      to the Agency.
                                  • D-78/VOA                            OLM04.2

-------
                                             Exhibit D -- Volatiles Appendix B
                  Modified SN-846 Method 5035 for Volatiles in Low Level Soils
9.0   PROCEDURE
9.1   The Contractor must determine whether a soil/sediment sample should be
      analyzed by the low or medium method.  Samples may contain higher than
      expected quantities of purgeable organics that will contaminate the
      purge-and-trap system thereby requiring extensive cleanup and instrument
      maintenance.  The Contractor may follow one of the screening procedures
      identified in Exhibit D-VOA Section 10.1.4.2.  The screening data are
      used to determine which is the appropriate sample preparation procedure
      for the particular sample.  If, based on the screening results, medium
      level analysis is required, the Contractor shall follow the procedure in
      Exhibit D-VOA Section 10.1.5.  If the Contractor received a pre-weighed
      sample preserved in methanol  (see Section 7.1.1), this sample shall be
      utilized for the medium level analysis.  It is the responsibility of the
      Contractor to analyze the sample at the correct level.

9.2   If insufficient sample amount  (less than 90% of the required amount) is
      received to perform the analyses, the Contractor shall contact SMO to
      notify them of the problem.  SMO will contact the Region for
      instructions.  The Region will either require that no sample analyses be
      performed or will require that a reduced volume be used for the sample
      analysis.

9.3   Sample Preparation

9.3.1    The following steps apply  to  the preparation of vials used for the
         analysis of low level soil/sediment samples by the closed-system
         purge-and-trap equipment described in this method.

9.3.2    Add a clean magnetic stirring bar to each clean vial.  If the purge-
         and-trap device employs a  means of stirring the sample other than a
         magnetic stirrer  (e.g., sonication or other mechanical means), then
         the stir bar is omitted.

9.3.3    Add approximately 1 g of sodium bisulfate preservative to each vial.
         If samples significantly smaller or larger than 5 g are to be used,
         adjust  the amount of preservative added to correspond to
         approximately 0.2 g of preservative for-each 1 g of sample.  Enough
         sodium bisulfate should be present  to ensure a sample pH of <2.

9.3.4    Add 5 ml/ of reagent water  to  each vial.   The water and the
         preservative will form an  acid solution that will reduce or eliminate
         the majority of the biological activity in the sample, thereby
         preventing biodegradation  of  the  volatile target compounds.

9.3.5    Seal  the vial with  the  screw-cap  and septum seal.  If the double-
         ended,  fritted vials are  used, seal both  ends as recommended by  the
         manufacturer.
                     ^
9.3.6    Affix a  label  to each  vial and weigh the prepared vial to the nearest
         0.01  g.  Record the tare  weight.

9.3.7    Because  volatile organics  will partition  into the headspace of the
         vial  from  the aqueous  solution and  will be lost when the vial is
         opened,  system monitoring compounds, matrix spikes and internal
         standards  should only be  added to the vials after  the sample has been

                                    D-79/VOA                            OLM04.2

-------
Exhibit D -- Volatiles Appendix B
Modified SW-846 Method 5035 for Volatiles in Low Level Soils

         added to the vial.  The standards should be introduced either
         manually by puncturing the septum with a small-gauge needle or
         automatically by the purge-and-trap system just prior to analysis.

9.3.8    Using the sample collection device, transfer the contents
          (approximately 5 g) into the sample vial containing the preservative
         solution.  This sample transfer must be performed rapidly to minimize
         loss of volatile compounds.  Quickly brush any soil off the vial and
         immediately seal the vial with the septum and screw-cap.  The soil
         vial is hermetically sealed and must remain so in order to guarantee
         the. integrity of the sample.  Gloves must be worn when handling the
         sample vial since the vial has been tared.  Record the date and time
         of  sample transfer onto the pre-prepared vials and submit with the
         data package.

         NOTE: Soil samples that contain carbonate minerals may effervesce
         upon contact with the acidic preservative solution in the sample
         vial.  Therefore, if samples are known or suspected to contain high
         levels of carbonates, a test sample  (from the 60 mL glass vial)
         should be added to a clean vial and checked for effervescence.  If a
         rapid or vigorous reaction occurs, the Contractor may discard the
         test sample and proceed with sample preparation by transferring the
         contents of the field core sampling/storage container into a clean
         vial that does not contain the preservative.

9.3.9    Weigh the vial and contents to the nearest 0.01 g and record this
         weight.  Sample weight is determined by subtracting the sample vial
         tared weight determined above from this final weight.

9.4   Sample Purge-and-Trap

5.4.1    Prior to sample purge, all soil/sediment samples must be allowed to
         warm to ambient temperature.  Shake the vial gently, to ensure that
         the contents move freely and that stirring will be effective.  Place
         the sample vial in the instrument carousel according to the
         manufacturer's instruction.

9.4.2    Without disturbing the hermetic seal on the sample vial, add 5 mL of
         reagent water, .10 uL of the internal standard spiking solution
          (Exhibit D-VOA Section 7.2.4.3), and 10 uL of the system monitoring
         compound .spiking solution  (Exhibit D-VOA Section 7.2.4.1).  All
         samples, including MS and USD, standards, and blanks, within an SDG
         must have the same amount of reagent water added.  Do not
         increase/change the amount of system monitoring compound and internal
         standard solution added.  Prior to purging, heat the sample vial to
         40  °C for 1.5 minutes,  or as described by the manufacturer.

9.4.3    Purge the sample with helium or another inert gas at a flow rate of
         20  to 40 mL/minute for 11 minutes while the sample is being agitated
         with the magnetic stirring bar or other mechanical means.  The purged
         analytes are allowed to flow out of the vial through a glass-lined
         transfer line to a trap packed with suitable sorbent materials.

9.4.4    If  a non-cryogenic interface is to be utilized, place the purge-and-
         trap system in the desorb mode after -the 11-minute purge, and preheat
         the trap to 180 °C without a flow of desorption gas.   Start the flow

                                   D-80/VOA                            OLM04.2

-------
                                             Exhibit D -- Volatiles Appendix B
                  Modified SW-846 Method 5035 for Volatiles in Low Level Soils

         of desorption gas at 10 mL/minute for about four minutes.   Begin the
         temperature program of the gas chromatograph and start data
         acquisition.

9.4.5    If a cryogenic interface is to be utilized, place the purge-and-trap
         system in the desorb mode after the 11-minute purge,  making sure that
         the cryogenic interface is at -150 °C or lower,  and rapidly heat  the
         trap to 180 °C while backflushing with an inert  gas at  4 mL/minute
         for about 5 minutes.  At the end of the 5-minute desorption cycle,
         rapidly heat the cryogenic trap to 250 °C.   Begin the temperature
         program of  the gas chromatograph and start the data acquisition.

9.4.6    After desorbing the sample for 4 to 5 minutes, recondition the trap
         by returning the purge-and-trap system to the purge mode..  Maintain
         the trap temperature at 180 °C.   After approximately  10 minutes,  turn
         off the trap heater and halt the purge flow through the trap.  When
         the trap is cool, the next sample can be analyzed.

9.5   Sample Dilutions

      If the on column concentration of any target compound exceeds the
      initial calibration range from the analysis of 5 g sample, a smaller
      sample size must be analyzed utilizing the procedure and methodology
      described in Exhibit D-VOA.  Guidance in performing dilutions and
      exceptions to  this requirement are given in Sections 10.1.6.2 through
      10.1.6.10 of Exhibit D-VOA.

9.6   Percent Moisture Determination

      It is highly recommended that the percent moisture determination only be
      made after the analyst has determined that no sample aliquots will be
      taken from the 60 mL vial for further analysis.  This is  to minimize
      loss of volatiles and  to avoid sample contamination from  the laboratory
      atmosphere.  The Contractor shall follow the procedure described in
      Exhibit D-VOA  Section  10.3 for determining percent moisture of  samples.

10.0  DATA ANALYSIS  AND CALCULATIONS

      The Contractor shall perform qualitative and  quantitative analysis  for
      the target and non-target compounds  following the procedures described
      in Exhibit D-VOA Section 11.0.  All  technical acceptance  criteria for
      sample analysis  described in Exhibit D-VOA  Section  11.3 shall be met or
      the corrective action  for sample analysis described in Section  11.4 of
      Exhibit D-VOA  shall be followed.

11.0  QUALITY CONTROL

11.1  Blank Analyses

11.1.1   Summary --  There are three different  types of blanks required by this
         method.

11.1.1.1    METHOD BLANK  -  a volume of purified solid  matrix  (prepared as
            described  in  Sections  9.3.2  through 9.3.5) and  carried  through  the
            entire analytical procedure.   If any samples are prepared without
             the  sodium bisulfate preservative,  a method  blank will be prepared

                                   D-81/VOA                            OLM04.2

-------
Exhibit D -- Volatiles Appendix B
Modified SW-846 Method 5035 for Volatiles in Low Level Soils

            in the same manner and run in the same 12 hour sequence as the
            unpreserved samples.  The weight of the purified solid matrix must
            be approximately equal to the weight of samples associated with
            the blank.  The purpose of a method blank is to determine the
            levels of contamination associated with the processing and
            analysis of samples.

11.1.1.2    STORAGE BLANK - upon receipt of the first samples in an SDG,  two
            of the sample vials to be used for the closed-system purge-and-
            trap analysis  (prepared as described in Sections 9.3.2 through
            9.3.5) are filled with reagent water.  The vials are stored with
            the samples in the SDG under the same conditions.  After all
            samples in the SDG have been analyzed, the storage blank is
            analyzed.  The storage blank indicates whether contamination may
            have occurred during storage of samples.

11.1.1.3    INSTRUMENT BLANK - a 5.0 mL aliquot of reagent water that is
            added to the sample vial  (prepared as described in Sections 9.3.2
            through 9.3.5) and carried through the entire analytical
            procedure.  Instrument blanks are analyzed after a sample which
            contains a target compound exceeding the initial calibration
            range.  The results from the instrument blank analysis indicate
            whether there is contamination from a previous sample.

11.1.2   Frequency of Blank Analyses

11.1.2.1    The method blank must be analyzed at least once during every
            12-hour time period on each GC/MS system used for volatile
            analysis  (see Section 9.2.2 of Exhibit D-VOA for the definition of
            the 12-hour time period).

11.1.2.2    The method blank must be analyzed after the continuing calibration
            and before any samples, including matrix spike/matrix spike
            duplicates, or storage blanks are analyzed.  The method blank must
            be analyzed after the initial calibration sequence if samples are
            analyzed before the 12-hour period expires.  A method blank must
            be analyzed in each 12-hour time period in which samples,
            including matrix spikes/matrix spike duplicates, and storage
            blanks are analyzed.

11.1.2.3    A minimum of one storage blank must be analyzed per SDG after all
            samples for that SDG have been analyzed.

11.1.2.4    The Contractor must demonstrate that there is no carryover from
            contaminated samples before data from subsequent, analyses may be
            used.  Samples may contain target compounds at levels exceeding
            the initial calibration range.  An instrument blank must be
            analyzed after the sample that exceeds the calibration range (also
            in the same purge inlet if an autosampler is used) or a sample
            that meets the maximum contamination criteria in Section 11.3.8 of
            Exhibit D-VOA must be analyzed.  For these purposes, if the
            instrument blank meets the technical acceptance criteria for blank
            analyses or the sample meets the maximum contamination criteria,
            the system is considered to be uncontaminated.  If the instrument
            blank or sample does not meet the criteria  (i.e., contaminated),
            the system must be decontaminated.  Until an instrument blank

                                   D-82/VOA                            OLM04.2

-------
                                             Exhibit D --  Volatiles Appendix B
                  Modified SW-846 Method 5035 for Volatiles in Low Level  Soils

            meets the blank technical acceptance criteria  or a sample meets
            the maximum contamination criteria in Section  11.3.8 of Exhibit D-
            VOA,  any samples analyzed since the original contaminated sample
            will require reanalysis at no additional cost  to the Agency.

         NOTE:  Only the instrument blank which demonstrates that there was no
         carryover from the previous sample or the instrument blank that
         demonstrates that the system is clean (Section 12.1.4 Exhibit D-VOA)
         needs to be reported.  Instrument blanks analyzed during the
         instrument decontamination process which exceed the requirements
         listed in Exhibit D-VOA Section 12.1.4 do not need to be reported.

11.1.3   Procedure for Blank Analyses

11.1.3.1    Method blanks shall be analyzed in the same manner as the
            associated samples, following the procedure described in Section
            9.4.

11.1.3.2    Storage/instrument blanks shall be analyzed in the same manner as
            the associated samples following the procedure outlined in section
            9.4.

11.1.3.3    A storage blank may be analyzed and reported as a soil sample if
          •  the SDG contains only soil samples.

11.1.3.4    Identify and quantitate analytes according to Section 11.0 of
            Exhibit D-VOA.

11.1.4   Technical Acceptance Criteria for Blank Analyses

11.1.4.1    All technical acceptance criteria for blank analyses described in
            Exhibit D-VOA Section 12.1.4 shall be met or corrective action for
            blank analyses described in Exhibit D-VOA Section 12.1.5 shall be
            followed.

11.2  Matrix  Spike/Matrix Spike Duplicate  (MS/MSD)

11.2.1   Summary of MS/MSD

         In order to evaluate  the effects of  the sample matrix on  the methods
         used for volatile analyses, the Agency has prescribed a mixture of
         volatile target compounds  to be spiked into  two aliquots of a sample,
         and  analyzed  in accordance  with the  appropriate method.

11.2.2   Frequency of  MS/MSD

11.2.2.1    A matrix spike and matrix spike duplicate must be performed for
            each group of samples of a  similar matrix for  the following,
            whichever  is most  frequent:

            •    Each SDG, or

            •    Each matrix within an SDG,  or

            •    Each group  of  samples of a  similar concentration  level
                   (soils only) ..  •

                                    D-83/VGA                            OLM04.2

-------
 Exhibit D -- Volatiles Appendix B
 Modified SW-846 Method 5035 for Volatiles in Low Level Soils

             MS/MSD samples shall be analyzed unless otherwise specified on the
             Traffic Report (TR) .  If no MS/MSD samples are specified on the
             TR, the Contractor shall contact SMO to confirm that MS/MSD
             analyses are not required.

 11.2.2.2    As a part of the Agency's QA/QC program,  water rinsate samples
             and/or field/trip blanks (field QC)  may accompany soil/sediment
             samples that are delivered to a laboratory for analysis.   The
             Contractor shall not perform MS/MSD analysis on any of the field
             QC samples.

 11.2.2.3    If the EPA Region designates a sample to be used as an MS/MSD,
             then that sample must be used.  If there is insufficient sample,
             less than the required amount, remaining to perform an MS/MSD,
             then the Contractor shall choose another sample to perform an
             MS/MSD analysis.  At the time the selection is made, the
             Contractor shall notify the Region (through SMO) that insufficient
             sample was received and identify the EPA sample selected for the
             MS/MSD analysis.  The rationale for the choice of a sample other
             than the one designated by the Region shall be documented in the
             SDG Narrative.

 11.2.2.4    If there is insufficient sample remaining for any of the samples
             in an SDG to perform an MS/MSD, then the Contractor shall
             immediately contact SMO to inform them of the problem.  SMO will
             contact the Region for instructions.   The Region will either
             approve that no MS/MSD is required,  or specify an alternative
             means of performing the MS/MSD analysis.   SMO will notify the
             Contractor of the resolution.  The Contractor shall document the
             decision in the SDG Narrative.

 11.2.2.5    If it appears that the Region has requested MS/MSD analysis at a
             greater frequency than required by the contract, the Contractor
             shall contact SMO.  SMO will contact the Region to determine which
             samples should have an MS/MSD performed on them.  SMO will notify
             the Contractor of the Region's decision.   The Contractor shall
             document the decision in the SDG Narrative.  If this procedure is
             not followed, the Contractor will not be paid for the MS/MSD
             analysis performed at a greater frequency than required by the
             contract.

 11.2.2.6    When a Contractor receives only a Performance Evaluation (PE)
             sample (s), no MS/MSD shall be performed within that SDG.

. 11.2.2.7    When a Contractor receives a PE sample as part of a larger SDG, a
             sample other than the PE sample must be chosen for the MS/MSD when
             the Region did not designate samples to be used for this purpose.

 11.2.3   Procedure for Preparing MS/MSD

 11.2.3.1    To prepare a matrix spike and matrix spike duplicate for low level
             soil/sediment samples, follow the procedure outlined in Section
             9.3.  Add 10 uL of the matrix spike solution (Exhibit D-VOA
             Section 7.2.4.2) either manually by puncturing the septum with a
             small-gauge needle or automatically by the purge-and-trap system
             just prior to analysis.  Analyze the matrix spike and matrix spike

                                    D-84/VOA                            OLM04.2

-------
                                             Exhibit D -- Volatiles Appendix B
                  Modified SW-846 Method 5035 for Volatiles in Low Level Soils

            duplicate samples by the procedure described in Section 9.4.  Do
            not further dilute MS/MSD samples to get either spiked or non-
            spiked analytes within calibration range.

11.2.4   Calculations for MS/MSD

         The Contractor shall calculate the concentrations of the matrix spike
         compounds in the matrix spike and matrix spike duplicate samples
         using the same equation as used for target compounds (Equation 6)
         Exhibit D-VOA Section 11.2.1.3.  The recovery of each matrix spike
         compound in the matrix spike and matrix spike duplicate samples and
         the relative percent difference  (RPD) of the recoveries shall be
         calculated as specified in Exhibit D-VOA Section 12.2.4.

11.2.5   Technical Acceptance Criteria for MS/MSD

         All technical acceptance criteria for MS/MSD specified in Exhibit D-
         VOA Section 12.2.5 must be met or corrective action for MS/MSD in
         Exhibit D-VOA Section 12.2.6 shall be followed.
                                    D-85/VOA                            OLM04.2

-------
     EXHIBIT D

ANALYTICAL METHODS
 FOR SEMIVOLATILES
     D-l/SVOA                             OLM04.2

-------
               Exhibit D  - Analytical Methods for SEMIVOLATILES

                               Table of Contents

Section                                                                     Page

1..0   SCOPE AND APPLICATION	   4

2.0   SUMMARY OF METHOD	 .  .   5
   2.1   Water   ........  	   5
   2.2   Low Level  Soil/Sediment   	^  5
   2.3   Medium  Level  Soil/Sediment  	   5

3.0   DEFINITIONS	   5

4.0   INTERFERENCES  	   6

5.0   SAFETY	   6

6.0   EQUIPMENT  AND  SUPPLIES   	   7

7.0   REAGENTS AND  STANDARDS	13
   7.1  . Reagents  ,	  13
   7.2   Standards   	13
   7.3   Storage of  Standard  Solutions	16

8.0   SAMPLE COLLECTION,  PRESERVATION, AND  STORAGE    	  18
   8.1   Sample  Collection  and Preservation	'	18
   8.2   Procedure  for Sample Storage	18
   8.3   Procedure  for Sample Extract Storage  	  18
   8.4   Contract Required  Holding Times    	  	  18

9.0   CALIBRATION AND  STANDARDIZATION	19
   9.1   Instrument Operating Conditions    	  19
   9.2   GC/MS Mass Calibration  (Tuning)  and Ion Abundance   	 19
   9.3   Initial Calibration   	  21
   9.4   Continuing Calibration	  .  23

10.0  PROCEDURE	.27
   10.1 Sample  Preparation	27
   10.2 Concentrating the  Extract	33
   10.3 Sample  Cleanup by  GPC	  35
   10.4 Sample  Extract Cleanup  by GPC    	40
   10.5 Final  Concentration   	  42
   10.6 Sample  Analysis by GC/MS	42

11.0  DATA ANALYSIS AND CALCULATIONS	44
   11.1 Qualitative Identification  	  44
   11.2 Calculations	47
   11.3 Technical  Acceptance Criteria  for Sample  Analysis   	 51
   11.4 Corrective Action    	  52

12.0  QUALITY CONTROL	55
   12.1 Method Blanks   	55
   12.2 Matrix Spike/Matrix Spike Duplicate (MS/MSD)   	 56


                                     D-2/SVOA                            OLM04.2

-------
13.0  METHOD PERFORMANCE   	  61

14.0  POLLUTION PREVENTION   	  61

15-. 0  -WASTE MANAGEMENT   	.- .	-.-;.-..-	61

16.0  REFERENCES   	61

17.0  TABLES/DIAGRAMS/FLOWCHARTS   .   . .	62

APPENDIX A - SCREENING OF SEMIVOLATILE ORGANIC EXTRACTS   	  71
   1.0   SCOPE AND APPLICATION   	71
   2.0   SUMMARY OF METHOD   	71
   3.0   INTERFERENCES	72
   4.0   SAFETY	72
   5.0   EQUIPMENT AND SUPPLIES	73
   6.0   REAGENTS AND STANDARDS	75
   7.0   QUALITY CONTROL   	76
   8.0   CALIBRATION AND STANDARDIZATION   	  77
   9.0   PROCEDURE	78
                                    D-3/SVOA                             OLM04.2

-------
Exhibit D Semivolatiles -- Section 1
Scope and Application

1.0   SCOPE AND APPLICATION

1.1   In 1978, EPA Headquarters and Regional representatives designed
      analytical methods  for  the analysis of semivolatiles in hazardous waste
      samples.  These methods were based on EPA Method  625, Base/Neutral and
      Acids.  In 1980,  these  methods were adopted for use in the  Contract
      Laboratory Program  (CLP).  As the requirements of CERCLA  evolved, the
      CLP methods, as well as their precedent EPA 600 Series methods,
      established the basis for other EPA methods to perform the  analysis of
      semivolatiles  contained in hazardous waste samples  (i.e., SW-846).  The
      following CLP  method has continuously improved to incorporate
      technological  advancements promulgated by EPA, and has continued to set
      the standard for  the preparation, extraction, isolation,  identification,
      and reporting  of  semivolatiles at hazardous waste sites.

1.2   The analytical method that follows is designed to analyze water,  soil
      and sediment from hazardous waste sites for the semivolatile organic
      compounds on the  Target Compound List  (TCL)  (see  Exhibit  C).   It covers
      the determination of a  number of organic compounds that are partitioned
      into  an organic solvent and are amenable to gas chromatography.   The
      method  involves solvent extraction of the matrix  sample,
      characterization  to determine the appropriate analytical  protocol to be
      used, followed by appropriate cleanup procedure and GC/MS analysis to
      determine the  semivolatile organic compounds present in the sample.

1.3    This  analytical method  provides  the  use of SH-846 Methods 3541
       (Revision 0, September  1994)  and 3545  (Revision  0, December 1996) for
       the extraction of soil/sediment samples.  However, prior  to using either
       one of  these alternate  extraction procedures, the Contractor must first
       demonstrate  that  these  procedures are equivalent  to  the existing
      procedures, and obtain  approval  for  use of these  alternative extraction.
      procedures  from  the EPA CLP National Program Manager.   The  process for
       determining and documenting  equivalency can be found in Exhibit  E,
       Section 6.0.

1.4    Problems have  been  associated with the following  compounds  analyzed by
       this  method.

1.4.1    Dichlorobenzidihe  and 4-chloroaniline  can be  subject  to  oxidative
          losses during  solvent concentration.

1.4.2    Hexachlorocyclopentadiene  is  subject  to thermal  decomposition in  the
          inlet  of the  gas chromatograph,  chemical  reactions  in  acetone
          solution,  and photochemical  decomposition.

 1.4.3    N-nitrosodiphenylamine decomposes in  the  gas  chromatograph  inlet
          forming  diphenylamine and,  consequently,  may  be  detected as
          diphenylamine.
                                     D-4/SVOA                            OLM04.2

-------
2.0
                                      Exhibit D Semivolatiles --  Sections 2  & 3
                                                  Summary of Method/Definitions
      SUMMARY OF METHOD
2.1   Water

      A one liter aliquot of sample is acidified to pH 2.0 and extracted with
      methylene chloride using a continuous liquid-liquid extractor.
      Separatory funnel extraction is NOT permitted.  The methylene chloride
      extract is dried with sodium sulfate, concentrated, subjected to GPC
      (GPC is required when higher molecular weight compounds are present that
      interfere with the analyses of target compounds; GPC is optional for all
      other circumstances), and analyzed by GC/MS for extractable organics.

2.2   Low Level Soil/Sediment

      A thirty  (30) gram portion of soil/sediment is mixed with anhydrous
      powdered sodium sulfate and extracted with 1:1 methylene
      chloride/acetone solution using an ultrasonic probe  or an alternate
      extraction method that has been previously accepted by the Agency for
      use by the Contractor.   If  the  low level  screen (Appendix A)  is used,  a
      portion of this dilute extract is concentrated fivefold and screened by
      GC/FID or GC/MS.  If peaks are present at greater than 10,000 ug/Kg,
      discard the extract and prepare the sample by the medium level method.
      If no peaks are present at greater than 10,000 ug/Kg, the entire extract
      is concentrated, subjected to GPC cleanup, and analyzed by GC/MS for
      extractable organics.

2.3   Medium Level  Soil/Sediment

      Approximately one gram portion of soil/sediment is mixed with anhydrous
      powdered sodium sulfate in a vial and extracted with methylene chloride.
      The methylene chloride extract can be screened for extractable organics
      using the procedure described in Appendix A, other EPA approved
      screening procedures or an in-house laboratory screening procedure.    If
      organic compounds are detected by the screen, the methylene chloride
      extract is subjected to GPC cleanup and analyzed by GC/MS for
      extractable organics.  If no organic compounds are detected by the
      medium level  screen, then a low level sample preparation is required.

3.0   DEFINITIONS

      See Exhibit G for a complete list of definitions.
                                    D-5/SVOA
OLM04.2

-------
Exhibit D Semivolatiles -- Sections 4 & 5
Interferences/Safety

4.0   INTERFERENCES

      Contaminants in solvents, reagents, glassware, and other sample
      processing hardware may cause method interferences such as discrete
      artifacts and/or elevated baselines in the extracted ion current
      profiles  (EICPs).  All of the materials routinely must be demonstrated
      to be free from interferences under the conditions of the analysis by
      running laboratory method blanks.  Matrix interferences may be caused by
      contaminants that are coextracted from the sample.  The extent of matrix
      interferences will vary considerably from source to source.

5.0   SAFETY

      The toxicity or carcinogenicity of each reagent used in this method has
      not been precisely determined; however, each  chemical should be treated
      as a potential health hazard.  Exposure to these reagents should be
      reduced to the lowest possible level.  The laboratory is responsible for
      maintaining a current awareness file of OSHA  regulations regarding the
      safe handling of the chemicals specified in this method.  A reference
     . file of data handling sheets  should be made available to all personnel
      involved  in these analyses.   Specifically, concentrated sulfuric acid
      presents,  some hazards and is  moderately toxic and extremely irritating
      to skin and mucous membranes.  Use these reagents in a fume hood
      whenever  possible and if eye  or skin contact  occurs flush with large
      volumes of water.  Always wear safety glasses or a shield for eye
      protection, protective  clothing, and observe  proper mixing when working
      with these reagents.
                                     D-6/SVOA                             OLM04.2

-------
                                           Exhibit D Semivolatiles -- Section 6
                                                         Equipment and Supplies
6.0   EQUIPMENT AND SUPPLIES
      Brand names, suppliers, catalog and part numbers are for illustrative
      purposes only.  No endorsement is implied.  Equivalent performance may
      be achieved using equipment and supplies other than those specified
      here, but demonstration of equivalent performance meeting the
      requirements of this Statement of Work is the responsibility of the
      Contractor.  The Contractor shall document any use of alternate
      equipment or supplies in the SDG Narrative.

6.1   Glassware

6.1.1    Continuous Liquid-Liquid Extractors - equipped with Teflon or glass
         connecting joints and stopcocks requiring no lubrication  (Hershberg-
         Wolf extractor, Ace Glass Company, Vineland, NJ P/N 6841-10 or
         equivalent) or Hydrophobic Membrane-based Extractor (Accelerated One
         Step™ Extractor, Corning series 3195 or equivalent).

6.1.2    Beakers  - 400 mL

6.1.3    Syringes - 2 uL, 10 uL, 0.2 mL, 0.5 mL, and 10 mL with Luerlok
         fitting.

6.1.4    Glass Scintillation Vials - at least 20 mL with screw-cap and Teflon
         or aluminum foil liner.

6.1.5    Pasteur  Pipets - 1 mL glass, disposable.

6.1.6    Vial and Caps - Amber glass, 2 mL capacity with Teflon-lined screw
         cap, 2 mL capacity for GC auto sampler.

6.1.7    Drying Column - 19 mm ID chromatographic column with coarse frit
          (substitution of a small pad of Pyrex glass wool for the frit will.
         help prevent cross contamination of sample extracts).

6.2   Kuderna-Danish  (K-D) Apparatus.

6.2.1    Concentrator Tubes - 15 mL and 10 mL graduated  (Kontes K-570050-1025
         or K-570040-1025 or equivalent).  Calibration must be checked at the
         volumes  employed in the test.  Ground-glass stoppers are used to
         prevent  evaporation of extracts.

6.2.2    Evaporative Flasks - 500 mL  (Kontes K-570001-0500 or equivalent).
         Attach to concentrator tube with springs.

6.2.3    Snyder Column - three-ball macro  (Kontes K-503000-0121 or
         equivalent).

6.2.4    Snyder Column - two-ball micro (Kontes K-569001-0219 or equivalent).

6.3   Spatula - stainless steel or Teflon.

6.4   Balances -  analytical, capable' of accurately weighing ± 0.0001 g, and
      one capable of weighing 100 g to ± 0.01 g.  The balances must be
      calibrated with class S weights or known reference weights once per each

                                    D-7/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 6
Equipment and Supplies

      12-hour work shift.  The balances must be calibrated with class S
      weights at a minimum of once per month.  The balances must also be
      annually checked by a certified technician.

6.5   Ultrasonic Cell Disrupters - Heat Systems, Ultrasonics Inc., Model W-385
      Sonicator  (475 watt with pulsing capability, No.200, M inch tapped
      disrupter horn, No. 419, 1/8 inch standard tapered Microtip probe, and .
      No. 305, 3/4 inch tapped high gain "Q" disruptor horn, or No. 208 3/4
      inch standard solid disruptor horn), or equivalent devices with a
      minimum of 375 watt output capability.  NOTE:  In order to ensure that
      sufficient energy is transferred to the sample during extraction, the
      Microtip probe or horn shall be replaced if the tip begins to erode.
      Erosion of the tip is evidenced by a rough surface.  .

6.6   Sonabox Acoustic Enclosure  (or equivalent) - for use with disruptor to
      decrease noise level.

6.7   Vacuum Filtration Apparatus.

6.7.1     Buchner Funnel.       •     '  •

6.7.2     Filter  Paper  - Whatman No. 41 or equivalent.

6J8   Pyrex Glass Wool  - rinsed with methylene chloride.

6.9   Test Tube Rack.

6.10  Silicon Carbide  Boiling  Chips - approximately  10/40 mesh.  Heat  to 400
      °C  for 30 minutes or Soxhlet extract with methylene chloride.  Teflon
      Boiling chips  solvent  rinsed prior to  use  are  acceptable.

6.11  Water Bath - heated, with concentric ring  cover, capable of  temperature
     • control  (±2  °C).   The  bath  should  be used in a hood.

6.12  Oven  - drying.

6.13  Desiccator.

6.14  Crucibles  - porcelain.

6.15  Nitrogen Evaporation Device  - equipped with a  water bath that  can be
      maintained at  35-40  °C.   (N-Evap by Organomation Associates, Inc.,  South
      Berlin, MA, or equivalent).   To prevent the  release of  solvent fumes
       into  the laboratory, the nitrogen  evaporator device must be  used in a
      hood.

6.16  pH Paper  -  including narrow range  capable  of measuring  a pH  of 2.0.

6.17  pH Meter  - with a combination glass  electrode,  calibrate according  to
      manufacturer's instructions.  The  pH meter shall be calibrated prior to
       each  use.
                                     D-8/SVOA                             OLM04.2

-------
                                           Exhibit D Semivolatiles -- Section 6
                                                         Equipment and Supplies
6.18  GPC Cleanup System
6.18.1   Gel Permeation Chromatography System - GPC Autoprep Model 1002 A or
         B, Analytical Biochemical Laboratories, Inc., or equivalent.  Systems
         that perform satisfactorily have been assembled from the following
         components - an HPLC pump, an auto sampler or a valving system with
         sample loops, and a fraction collector.  All systems, whether
         automated or manual, must meet the calibration requirements of
         Section 10.3.3.

6.18.2   NOTE:  GPC cleanup is  required for all soil/sediment extracts,  and
         for water extracts containing higher molecular weight contaminants
         that interfere with the analyses of the target compounds.

6.18.3   Chromatographic Column - 700 mm x 25 mm ID glass column.  Flow is
         upward.  To simplify switching from the UV detector during
         calibration to the GPC collection device during extract cleanup, an
         optional double 3-way valve  (Rheodyne Type 50 Teflon Rotary Valve
         #10-262 or equivalent) may be attached so that the column exit flow
         can be shunted either to the UV flow-through cell or to the GPC
         collection device.

6.18.4   Guard Column  (optional) - 5 cm, with appropriate fittings to connect
         the inlet side of the analytical column (Supelco 5-8319 or
         equivalent) ^

6.18.5   Bio Beads  (SX-3) - 200-400 mesh, 70 g  (Bio-Rad Laboratories,
         Richmond, CA, Catalog 152-2750 or equivalent).  An additional 5 g of
         Bio Beads are required if the optional guard column is employed.  The
         quality of Bio Beads may vary from lot to lot because of excessive
         fines in some lots.  In addition to fines having a detrimental effect
         on Chromatography, they can also pass through the column screens and
         damage the valve.

6.18.6   Ultraviolet Detector - fixed wavelength (254 nm) with a semi-prep
         flow-through cell.

6.18.7   Strip chart recorder, recording integrator, or laboratory data
         system.

6.18.8   Syringe Filter Assembly disposable - Bio-Rad "Prep Disc" sample
         filter assembly #343-0005, 25 mm, and 5 micron filter discs or
         equivalent.  Note:  Some instrument manufacturers recommend a smaller
         micron size filter disc.  Consult the instrument operation manual to
         determine the proper filter disc to use in the system.  Check each
         batch for contaminants.  Rinse each filter assembly  (prior to use)
         with methylene chloride if necessary.

6.19  Gas Chromatograph/Mass Spectrometer (GC/MS) System

6.19.1   Gas Chromatograph - The-gas chromatograph  (GC) system must be capable
         of temperature programming and have a flow controller that maintains
         a constant column flow rate throughout the temperature program
         operations.  The system must be suitable for splitless injection and
         have all required accessories including syringes, analytical columns,
                                    D-9/SVOA
OLM04.2

-------
Exhibit D Semivolatiles -- Section 6
Equipment and Supplies

         and gases.  All GC carrier gas lines must be constructed from
         stainless steel or copper tubing.  Non-polytetrafluoroethylene  (PTFE)
         thread sealants or flow controllers with rubber components are not to
         be used.

6.19.2   Gas Chromatography Column - Minimum length 30 m x  0.25 mm ID  (or 0.32
         mm) bonded-phase  silicon coated fused silica capillary column DB-5
          (J&W Scientific); RTX-5  (Restek); SPB-5  (Supelco); AT-5  (Alltech);
         HP-5  (Hewlett-Packard); CP-Sil 8CB  (Chrompack); 007-2  (Quadrexj ; BP-5
          (SGE); or equivalent.  Note that this is a minimum requirement  for
         column length.  Longer columns may be used.  Although a  film
         thickness of  1.0  micron is recommended because of  its larger
         capacity, a film  thickness of 0.25 micron may be used.  A description
         of the GC column  used for analysis shall be provided in  the SDG
         Narrative.

6.19.2.1       A  capillary column  is  considered  equivalent  if:

               •   The  column does  not introduce  contaminants  which interfere
                   with the identification  and  quantification  of  the  compounds
                   listed in Exhibit C (Semivolatiles).

               •   The  analytical  results  generated using the  column  meet the
                   initial and continuing  calibration technical  acceptance
                   criteria listed in the  SOW,  and the CRQLs listed in Exhibit
                   C (Semivolatiles) .

               •   The  column can accept  up to  160 ng of  each compound listed
                   in Exhibit C (Semivolatiles)  without becoming overloaded.

                •   The  column provides equal or better resolution of  the
                   compounds listed in Exhibit  C (Semivolatiles)  than columns
                   listed in Section 6.19.2.

6.19.2.2       As applicable, follow manufacturer's instructions for use of
                its product.

6.19.2.3        The Contractor must maintain documentation that the column met
                the criteria in Section 6.19.2.1.   The minimum documentation  is
                as follows:

6.19.2.3.1      Manufacturer provided information concerning the performance
                characteristics of the column.

6.19.2.3.2      Reconstructed ion chromatograms and data system reports
                generated on the GC/MS used for CLP analyses:

                •  From blanks which demonstrate that there are no contaminants
                   which interfere with the semivolatile analysis when using
                   the column;

                •  For initial calibration standards.analyzed using the column;

                •  For continuing calibration standards analyzed using the
                   column.

                                    D-10/SVOA                            OLM04.2

-------
                                           Exhibit D Semivolatiles -- Section 6
                                                         Equipment and Supplies

6.19.2.3.3     Based on the Contractor generated data described in Section
               6.19.2.3.2, the Contractor must complete a written review,
               signed by the Laboratory Manager certifying that:

               •  The column performance meets the  technical acceptance
                  criteria in Sections 9.3.5 and  9.4.5.

               •  The low point  initial calibration standard analysis has
                  adequate sensitivity to meet the  semivolatile CRQLs.

               •  The high point  initial calibration standard analysis was not
                  overloaded.

               •  The column does not introduce contaminants which interfere
                  with the identification and/or  quantitation of  compounds
                  listed in Exhibit  C  (Semivolatiles).

6.19.2.4       The documentation  must be made available to the Agency during
               on-site laboratory evaluations or  sent to the Agency  upon
               request of the Technical Project Officer or the Administrative
               Project Officer.

6.19.2.5       Packed columns cannot be used.

6.19.3   Mass Spectrometer - must be capable of scanning from 35 to  500 amu
         every 1 second or less,  utilizing 70 volts  (nominal) electron energy
         in the electron impact ionization mode, and producing a mass spectrum
         which meets the tuning acceptance criteria  when 50 ng of
         decafluorotriphenylphosphine  (DFTPP) is injected through the GC
         inlet.  The instrument must be vented to the outside of the facility
         or to a trapping system  which prevents the  release of contaminants
         into the instrument room.

6.19.4   GC/MS interface - any gas chromatograph to  mass spectrometer
         interface which provides acceptable sensitivity at contract required
         quantitation limits.  Gas chromatograph to  mass spectrometer
         interfaces constructed of all-glass or glass-lined materials are
         recommended.  Glass can  be  deactivated by  silanizing with
         dichlorodimethylsilane.

6.19.5   Data system - a computer system interfaced  to the mass spectrometer
         that allows the continuous  acquisition and  storage on machine-
         readable media of all mass  spectra obtained throughout the  duration
         of the chromatographic program.  The computer must have
         software that allows searching any GC/MS data file for ions of a
         specified mass and plotting such ion abundance versus time  or scan
         number.  This type of plot  is defined as an Extracted Ion Current
         Profile  (EICP).  Software must also be available that allows
         integrating the abundance in any EICP between specified time or scan
         number limits.  Also, for the non-target compounds, software must be
         available that allows for the comparison of sample spectra  against
         reference library spectra.  The NIST/EPA/NIH (May 1992 release or
         later)and/or Wiley (1991 release or later)  or equivalent mass
         spectral library shall be used as the reference library.  The
         operational data system  must be able to flag the data files that have

                                   D-11/SVOA                             OLM04.2

-------
Exhibit D Semivblatiles -- Section 6
Equipment and Supplies

         been edited manually by laboratory personnel.

6.19.6   Magnetic tape storage device - must be capable of recording data and
         suitable for long-term, off-line storage of GC/MS data.
                                    D-12/SVOA                            OLM04.2

-------
7.0
                                           Exhibit D Semivolatiles -- Section 7
                                                         Reagents and Standards
REAGENTS AND STANDARDS
7.1   Reagents

7.1.1    Reagent Water - defined as water in which an interferant is not
         observed at or above the CRQL for each analyte of interest.  Reagent
         water may be generated by passing tap water through a carbon filter
         bed containing about 453 g of activated carbon  (Calgon Corp.,
         Filtrasorb - 300 or equivalent).

7.1.2    Sodium Thiosulfate -  (ACS) granular.

7.1.3    Sulfuric Acid Solution  (H2SO4)  - (1+1) slowly add 50 mL of
         concentrated H2S04 (sp.  gr.  1.84;  18 N)  to 50 mL of reagent water.

7.1.4    Acetone, methanol, methylene chloride, iso-octane, 2-propanol, and
         toluene - pesticide residue analysis grade or equivalent.

7.1.5    Sodium Sulfate - powdered or granular anhydrous reagent grade, heated
         at 400  °C for four hours in a shallow tray, cooled in a desiccator,
         and stored in a glass bottle (Baker anhydrous powder; catalog #73898;
         Baker anhydrous granulated, catalog #3375; or equivalent).  Caution:
         An open container of sodium sulfate may become contaminated during
         storage in the laboratory.

7.1.6    Sodium Sulfite - reagent grade.

7.2   Standards
7.2.1
         Introduction
7.2.2
   The Contractor must provide all standards to be used with this
   contract.  These standards may be used only after they have been
   certified according to the procedure in Exhibit E.  The Contractor
   must be able  to verify that the standards are certified.
   Manufacturer's certificates of analysis must be retained by the
   Contractor and presented upon request.

   Stock Standard Solutions
7.2.2.1     Stock  standard  solution may be purchased or prepared using the
            following procedure.

7.2.2.1.1       Prepare  stock  standard  solutions  by  accurately weighing about
                0.0100 g of  pure material.  Dissolve the material in methylene
                chloride or  another suitable  solvent and dilute  to volume  in a
                10  mL volumetric flask.  Larger volumes may be used at the
                convenience  of the analyst.

7.2.2.1.2       When compound  purity is assayed to be 97.0 percent or greater,
                the weight may be used without correction to  calculate the
                concentration  of the stock solution.  If the  compound purity is
                assayed  to be  less than 97.0  percent, the weight must be
                corrected when calculating the concentration  of  the stock
                solution.  See Exhibit E  (Analytical Standards Requirements).
                                   D-13/SVGA
                                                                   OLM04.2

-------
Exhibit D Semivolatiles  --  Section 7
Reagents and Standards

7.2.2.1.3       Fresh stock standards  must  be  prepared once every twelve
                months,  or sooner if standards have degraded or concentrated.
                Stock standards must be checked for signs of degradation  or
                concentration just prior to preparing secondary dilution  and
                working standards from them.

7.2.3     Secondary Dilution Standards
                                              (
7.2.3.1      Using stock standards, prepare secondary dilution standards  in
             methylene chloride that contain the compounds of interest  either
             singly or mixed together.

7.2.3.2      Fresh secondary dilution  standards must be prepared once every
             twelve months, .or sooner  if standards have degraded or
             concentrated.  Secondary  dilution standards must be checked  for
             signs of degradation or concentration just prior to preparing
             working standards from them.

7.2.4     Working Standards

7.2.4.1      Surrogate Standard Spiking Solution

             Prepare a surrogate standard spiking solution that contains
             nitrobenzene-d5, terphenyl-d!4,  2-fluorobiphenyl, and 1,2-
             dichlorobenzene-d4 at a concentration of 100 ug/mL; phenol-d5,
             2,4, 6-tribromophenol, 2-fluorophenol, and 2-chlorophenol-d4  at a
             concentration of 150 ug/mL.  Surrogate standards are added to all
             samples and calibration solutions.  Additional surrogates  may be
             added at the laboratory's discretion.

•7.2.4.2      Matrix Spiking Solution

 7.2.4.2.1      The matrix  spiking solution consists of the following:

                Bases/Neutrals                    Acids
                Acenaphthene                      Pentachlorophenol
                2,4-Dinitrotoluene                Phenol
                Pyrene                            2-Chlorophenol
                N-Nitroso-di-n-propylamine        4-Chloro-3-methylphenol
                                                  4-Nitrophenoi

 7.2.4.2.2      Prepare  a  spiking solution that contains each of the
                base/neutral  compounds above at 100 ug/mL  in methanol and the
                acid  compounds at 150 pg/mL in methanol.

 7.2.4.3      GPC Calibration  and GPC  Continuing Calibration Verification
             Solution
                                    D-14/SVOA                            OLM04.2

-------
                                           Exhibit D Semivolatiles -- Section 7
                                                         Reagents and Standards

7.2.4.3.1      Prepare a calibration solution in methylene chloride  containing
               the following analytes at the minimum concentration listed  (in
               elution order):

               Compound             Concentration  (mg/mL)
               Corn oil                                25.0
               bis(2-ethylhexyl)phthalate               O.5
               Methoxychlor                             0.1
               Perylene                                 0.02
               Sulfur                                   0.08

7.2.4.3.2      NOTE:  Sulfur is not very soluble in methylene chloride, but  it
               is soluble in warm corn oil.  Therefore, one approach is to
               weigh out the corn oil, warm it, and transfer the weighed
               amount of sulfur into the warm corn oil.  Mix it and  then
               transfer into a volumetric  flask with methylene chloride, along
               with the other calibration  compounds.

7.2.4.4     Instrument Performance Check Solution

               Prepare a solution of decafluorotriphenylphosphine  (DFTPP),
               such that a 2 uL injection will contain 50 ng of DFTPP.  The
               DFTPP may also be included  in the calibration standards at this
               level.

7.2.4.5     Initial and Continuing Calibration Solutions

7.2.4.5.1      Prepare calibration standards at a minimum of five
               concentration levels  (20, 50, 80, 120, and 160 total  ng per 2
               uL).  Each calibration standard should contain each compound  of
               interest and each surrogate.  Eight compounds  (2,4-
               Dinitrophenol, 2,4,5-Trichlorophenol, 2-Nitroaniline, 3-
               Nitroaniline, 4-Nitroaniline, 4-Nitrophenol, 4,6-Dinitro-2-
               methylphenol, and Pentachlorophenol) will require only a four-
               point initial calibration at 50, 80, 120, and 160 total ng per
               2 uL since detection at less than 50 ng per injection is
               difficult.

7.2.4.5.2      In order to facilitate the  confirmation of single component
               pesticides from the semivolatile library search data  (see
               Exhibit D  (Pesticides/Aroclors), Section 12.1.2), the
               laboratory may include the  single component pesticide target
               compounds listed in Exhibit C in the semivolatile continuing
               calibration standard.  The  laboratory may add any or  all of
               these compounds to the semivolatile continuing calibration
               standard, but at a concentration of 10 ng/uL or less.  Do not
               include the Aroclors or toxaphene mixtures in the semivolatile
               initial and continuing calibration standards.  If added to this
               GC/MS standard, these additional analytes are not reported on
               the semivolatile calibration form  (Form VII), but must be
               included in the quantitation report for the continuing
               calibration standard.  As only a single point calibration would
               be performed, no %RSD or percent difference criteria  would
               apply to these additional analytes.
                                   .D-15/SVOA                             OLM04.2

-------
 Exhibit D  Semiyolatiles  --  Section  7
 Reagents and  Standards

 7.2.4.5.3       The 50 ng/2 uL calibration standard is the continuing
                 calibration standard.

 7.2.4.6      Internal Standard Solution

              An internal standard solution can be prepared by dissolving 100 mg
              of each of the following compounds in 15 mL of methylene chloride:
              1,4-dichlorobenzene-d4, naphthalene-d8,  acenaphthene-dlO,
              phenanthrene-dlO, chrysene-d!2 and perylene-d!2.   It may be
              necessary to use 5.0 to 10.0 percent toluene in this solution and
              a few minutes of ultrasonic mixing in order to dissolve all the
              constituents.  The resulting solution will contain each standard
              at a concentration of 2000 ng/uL.  A 10 uL portion of this
              solution should be added to each 1 mL of sample extract just prior
              to analysis by GC/MS.  This will result in 40 ng of each internal
              standard in the 2 uL volume of extract injected into the GC/MS.

              NOTE:  For automated systems using an injection volume of less
              than 10 uL, the internal standard solution may need to be prepared
              at a different concentration.  Prepare the internal standard
              solution such that the aliquot used by the system maintains the
              required 40 ng of each internal standard in the 2 uL volume of
              extract injected into the GC/MS.

  7.2.5    Ampulated Standard Extracts

           Standard solutions purchased from a chemical supply house as
           ampulated extracts in glass vials may be retained and used until the
           expiration date provided by the manufacturer.  If no manufacturer's
           expiration date is provided, the standard solutions as ampulated
           extracts may be retained and used for 2 years from the preparation
           date.  Standard solutions prepared by the Contractor which are
           immediately ampulated in glass vials may be retained for 2 years from
           the preparation date.  Upon breaking the glass seal, the expiration
           times listed in Sections 7.2.2.1.3, 7.2.3.2, and 7.3.3 will apply.
           The Contractor is responsible for assuring that the integrity of the
           standards have not degraded  (see Section 7.3.5).

  7.3   Storage of Standard Solutions

  7.3.1    Store the stock and secondary dilution standard solutions at
|           4°C  (± 2°C) in Teflon-lined screw-cap amber bottles.  Fresh standards
           should be prepared every twelve months at a minimum.

|  7.3.2    Store the working standards at  4°C (± 2°C)  in Teflon-sealed
           containers.  The  solution should be checked frequently for stability.
           These solutions must be replaced after twelve months, or sooner if ,
           comparison with quality control check samples indicates a problem.

  7.3.3    The  continuing  calibration standard  (50 ng) should be prepared weekly
|           and  stored at  4 °C (±  2°C).   Refrigeration  of  the  GPC  calibration
           solution may cause  the corn oil to precipitate.  Before use, allow
           the  solution to stand at  room temperature until the corn oil
           dissolves.  Replace this  calibration solution every six months, or
           more  frequently if  necessary.

                                   .  D-16/SVOA                             OLM04.2

-------
                                           Exhibit D Semivolatiles -- Section 7
                                                         Reagents and Standards

7.3.4    Protect all standards from light.  Samples, sample extracts, and
         standards must be stored separately.

7.3.5    The Contractor is responsible for maintaining the integrity of
         standard solutions and verifying prior to use.  Storage of standard
         solutions in the freezer may cause some compounds to precipitate.
         This means, at a minimum, the standards must be brought to room
         temperature prior to use, checked for losses, and checked that all
         components have remained in solution.  Additional steps may be
         necessary to ensure all components are in solution.
                                   D-17/SVOA                            OLM04.2

-------
  Exhibit  D Semivolatiles --  Section 8
  Sample Collection,  Preservation,  and  Storage

  8.0    SAMPLE COLLECTION, PRESERVATION,  AND STORAGE

  8.1    Sample.Collection and Preservation

  8.1.1    Water samples may be collected in 1 L (or 1 quart)  amber glass
           containers, fitted with screw-caps lined with Teflon.  If amber
           containers are not available,  the samples should be protected from
           light.  Soil samples may be collected in glass containers or closed
           end tubes  (e.g., brass sleeves) in sufficient quantity to perform the
           analysis.  The specific requirements for site.sample collection are
           outlined by the Region.

  8.1.2    All samples must be iced or refrigerated at 4  °C (±2 °C) from the
           time of collection until extraction.

  8.2    Procedure for Sample Storage

  8.2.1    The samples must be protected from light and refrigerated at
           4  °C  (±2 °C) from the time of receipt until 60 days after delivery of
           a  complete, reconciled data package to the Agency.  After 60 days the
           samples may be disposed of in a manner that complies with all
           applicable regulations.

  8.2.2    The samples must be stored in an atmosphere demonstrated to be free
           of all potential contaminants.

  8.3    Procedure for Sample Extract Storage

|  8.3.1    Sample extracts must be protected  from light and stored  at   4°C
|            (± 2°C) until  365 days after delivery of a reconciled, complete data
           package to the Agency.

  8.3.2    Samples, sample extracts, and  standards must be stored separately.

  8.4   Contract Required Holding Times

  8.4.1    Extraction of  water-samples by continuous  liquid-liquid  procedures
           shall be started within  5 days of  Validated Time of  Sample Receipt
            (VTSR).  Extraction of soil/sediment samples by sonication procedures
           shall be completed within 10 days  of VTSR.  NOTE:  Separatory  funnel
           extraction procedures are not  permitted.

  8.4.2    As part of the Agency's  QA program, the Agency may provide
           Performance  Evaluation  (PE) samples as standard extracts which the
           Contractor is required to prepare  per the  instructions provided by
|           the Agency.   PE samples must  be  prepared and analyzed concurrently
|           with  the  samples  in  the  SDG.   The extraction holding time (5 days
           after VTSR for water  and 10 days after VTSR  for soil/sediment) does
           not apply  to PE samples  received as standard  extracts.

  8.4.3    Extracts of  water and soil/sediment samples must be  analyzed within
           40 days  following extraction.
                                      D-18/SVOA                            OLM04.2

-------
                                           Exhibit D Semivolatiles -- Section 9
                                                Calibration and Standardization
           Instrument Operating Conditions/GC/MS Calibration and Ion Abundance

9.0   CALIBRATION AND STANDARDIZATION

9.1   Instrument Operating Conditions

9.1.1    Gas Chromatograph

9.1.1.1     The following are the gas chromatographic analytical condition.
            The conditions are recommended unless otherwise noted.

            Initial Column Temperature Hold      40 °C for 4 minutes
            Column Temperature Program           40-270 °C at 10 °C/min.
            Final Column Temperature Hold        270 °C;
                                                 Hold Required: 3 minutes
                                                 after all compounds listed
                                                 in Exhibit C  (SVOA). have
                                                 eluted
            Injector Temperature                 250-300  °C
            Transfer Line Temperature            250-300  °C
            Source Temperature                   According to manufacturer's
                                                 specifications
            Injector                             Grob-type, splitless
            Sample Volume                        2 uL
            Carrier Gas                          Helium at 30 cm/sec

9.1.1.2     Optimize GC conditions  for analyte separation and  sensitivity.
            Once optimized,  the same GC  conditions must be used for  the
            analysis of all  standards, samples, blanks, matrix spikes, and
            matrix spike duplicates.

9.1.2    Mass Spectrometer

         The following are the required  mass spectrometer analytical
         conditions:

            Electron Energy                      70 volts  (nominal)
            Mass Range                           35 to 500 emu
            Scan Time                            Not to exceed 1 second per
                                                 scan

9.2   GC/MS Mass Calibration (Tuning) and Ion Abundance   •

9.2.1    Summary of GC/MS Instrument Performance Check

         The GC/MS system must be tuned  to meet the manufacturer's
         specifications, using a suitable calibration such as perfluoro-tri-n-
         butylamine  (FC-43)  or perfluorokerosene  (PFK).   The mass calibration
         and resolution of the GC/MS system are verified  by the analysis of
         the instrument performance check solution  (Section 7.2.4.4).  Prior
         to the analysis of  any samples, including matrix spike/matrix spike
         duplicates, blanks, or calibration standards, the Contractor must
         establish that the  GC/MS system meets the mass spectral ion abundance
         criteria for the instrument performance check solution containing
         decafluorotriphenylphosphine  (DFTPP).
                                    D-19/SVOA                             OLM04.2

-------
Exhibit D Semivolatiles -- Section 9
Calibration and Standardization
GC/MS Calibration and Ion Abundance

9.2.2    Frequency of GC/MS  Instrument Performance Check

9.2.2.1     The instrument performance check  solution must be analyzed once at
            the beginning of each  12-hour period during which samples or
            standards are analyzed.

9.2.2.2     The 12-hour time period for  a GC/MS system instrument performance
            check  and standards  calibration  (initial or continuing  calibration
            criteria) begins at  the moment of injection of the DFTPP analysis
            that the laboratory  submits  as documentation of  a compliant
            instrument performance check.  The time period ends  after 12  hours
            have elapsed according to the system clock.

9.2.3    Procedure  for GC/MS Instrument  Performance Check

         The  analysis of the instrument  performance check solution  may be
         performed as an injection of up to  50 ng of DFTPP into  the GC/MS or
         by adding 50 ng of  DFTPP  to  the calibration standards  (Section
         7.2.4.5)  and analyzing  the calibration standard.

9.2.4    Technical  Acceptance Criteria  for GC/MS Instrument  Performance Check

9.2.4.1     The GC/MS  system must  be  tuned at the  frequency  described in
            Section 9.2.2.

9.2.4.2     The abundance  criteria listed  in Table  1 must be met for a  50 ng
             injection  of  DFTPP.   The  mass  spectrum of  DFTPP  must be acquired
             in the following manner:   Three  scans  (the peak  apex scan and the
             scans  immediately preceding and  following  the apex)  are acquired
             and averaged.   Background subtraction  is  required,  and must be
             accomplished using a single scan acquired  no more  than 20 scans
            prior  to the  elution of  DFTPP.   Do not subtract  part of the DFTPP'
            peak.   NOTE:   All subsequent standards,  samples, MS/MSD, and
             blanks associated with a DFTPP analysis must use the identical
            mass  spectrometer instrument conditions.

9.2.5     Corrective Action for GC/MS  Instrument  Performance  Check

9.2.5.1      If the DFTPP acceptance criteria are not met,  re-tune the GC/MS
             system.  It may be necessary to clean the  ion  source,  clean
             quadrupoles,  or take other actions to achieve  the  technical
             acceptance criteria.

9.2.5.2      DFTPP acceptance criteria MUST  be met  before any standards,
             samples, including MS/MSD,  or required blanks  are  analyzed.   Any
             standards,  samples,  or required blanks analyzed when tuning
             technical acceptance criteria have not been met  will require
             reanalysis at no additional cost to the Agency.
                                    D-20/SVOA                            OLM04.2

-------
                                           Exhibit D Semivolatiles -- Section 9
                                                Calibration and Standardization
                                                            Initial  Calibration

9.3   Initial Calibration

9.3.1    Summary of Initial Calibration      ., _.   .      __..._.

         Prior to the analysis of samples and required blanks, and after the
         instrument performance check solution criteria have been met, each
        • GC/MS system must be calibrated at a minimum of five concentrations
         (Section 7.2.4.5) to determine instrument sensitivity and the
         linearity of GC/MS response for the semivolatile target and surrogate
         compounds.

9.3.2    Frequency of Initial Calibration

9.3.2.1     Each GC/MS system must be initially calibrated>upon award of the
            contract, whenever the Contractor takes corrective action which
            may change or affect the initial calibration criteria  (e.g., ion
            source cleaning or repairs, column replacement, etc.), or if the
            continuing calibration technical acceptance criteria  have not been
            met.

9.3.2.2     If time still remains in the 12-hour  time period after meeting the
            technical acceptance criteria for the initial calibration, samples
            may be analyzed.  It is not necessary to analyze a continuing
            calibration standard within this 12-hour time period, if the
            initial calibration standard that is  the same concentration as the
            continuing calibration standard meets the continuing  calibration
            technical acceptance criteria.  Quantitate all sample and quality
            control sample results and quality control criteria results, such
            as internal standard area response change and retention  time
            shift, against the initial calibration standard that  is  the same
            concentration as the continuing calibration standard  (50 ng/2 uL).

9.3.3    Procedure for Initial Calibration

9.3.3.1     All standard/spiking solutions and blanks must be allowed to warm
            to ambient temperature  (approximately 1 hour) before  preparation
            or analysis.

9.3.3.2     Prepare five calibration standards containing all the semivolatile
            target and surrogate compounds at the concentrations  described in
            Section 7.2.4.5.

9.3.3.3     Add a 10 uL aliquot of internal standard solution  (Section
            7.2.4.6) to 1 tnL aliquot of calibration standards to  result in 40
            ng of internal standard in the 2 uL volume of calibration standard
            injected onto the GC/MS.  The internal standards specified in
            Section 7.2.4.6 should permit most of the semivolatile target
            compounds to have relative retention  times of 0.80 to 1.20, using
            the assignments of internal standards to target compounds given in
            Table 2.

9.3.3.4     Analyze each calibration standard by  injecting 2.0 uL of standard.
                                   D-21/SVOA                             OLM04.2

-------
Exhibit D Semivolatiles -- Section 9   -^
Calibration and Standardization
Initial Calibration

9.3.4    Calculations  for Initial Calibration

9.3.4.1     Calculate  the relative response  factors  (RRF)  for  each
            semivolatile target and  surrogate  compound using Equation 1  and
            the primary characteristic  ions  found  in  Table 3  (Internal
            Standards) and Table 4  (Target Compounds  and  Surrogates).  Assign
            the target compounds and surrogate to  the internal standard
            according  to Table 2.  For  internal standards,  use the  primary ion
            listed in  Table  3 unless interferences are present.   If
            interferences prevent the use of the primary  ion for a  given
            internal standards, use  the secondary  ion(s)  listed in  Table 3;

            NOTE:   Unless otherwise  stated the area  response of the primary
            characteristic ion is the quantitation ion.
             EQ.
                                 RRF =     x    !
                                       A.    C
             Where,

             Ax  =  Area of the characteristic ion for the compound to be
                   measured (see Table 4) .
             Ais =  Area of the characteristic ion for specific internal
                   standard (see Table 3) .
             Cis =  Amount of the internal standard injected (ng).
             Cx  =  Amount of the compound to be measured injected (ng).

 9.3.4.2      The mean relative response factor (RRF)  must be calculated for all
             compounds.  Calculate the % Relative Standard Deviation  (%RSD) of
             the RRF values for the initial calibration using the  following
             equation:

             EQ.  2
                                Standard Deviation
                         %RSD =
                                       Mean
             Where,
                                              £  (x/-x):
                       Standard Deviation =
             Xt = each individual value used to calculate the mean.
             x  = the mean of n values.
             n  = the total number of values.
                                    D-22/SVOA                             OLM04.2

-------
                                           Exhibit D Semivolatiles -- Section 9
                                                Calibration and Standardization
                                                         Continuing Calibration

9.3.5    Technical Acceptance Criteria for Initial Calibration

9.3.5.1     All initial calibration standards must be analyzed at the
            concentration levels described in Section 7.2.4.5 and at the
            frequency described in Section 9.3.2 on a GC/MS system meeting the
            DFTPP technical acceptance criteria.

9.3.5.2     The relative response factor  (RRF) at each calibration
            concentration for each semivolatile target and surrogate compound
            must be greater than or equal to the compound's minimum acceptable
            relative response factor listed in Table 5.

9.3.5.3     The %RSD over the initial calibration range for relative response
            factor for each semivolatile and surrogate compound that has a
            required %RSD must be less than or equal to the %RSD listed in
            Table 5.

9.3.5.4     Up to four compounds may fail to meet the criteria listed in Table
            5.  However, these four compounds must meet a minimum RRF
            criterion of 0.010 and have a %RSD less than or equal to 40.0
            percent.

9.3.5.5     Excluding those ions in the solvent front, no quantitation ion may
            saturate the detector, with the exception of quantitation ions in
            up to 3 compounds  (including internal standards and surrogates)
            from the high standard  (160 ng/2 pL).  Consult the manufacturer's
            instrument manual to determine how saturation is indicated for
            your instrument.

9.3.6    Corrective Action for Initial Calibration

9.3.6.1     If any technical acceptance criteria for initial calibration are
            not met, inspect the system for problems.  It may be necessary to
            clean the ion source, change the column, or take other corrective
            actions to achieve the acceptance criteria.

9.3.6.2     Initial calibration technical acceptance criteria must be met
            before any samples, including matrix spike/matrix spike duplicates
            or required blanks, are analyzed.  Any samples, including matrix
            spike/matrix spike duplicates or required blanks, analyzed when
            initial calibration technical acceptance criteria have not been
            met will require reanalysis at no additional cost to the Agency.

9.4   Continuing Calibration

9.4.1    Summary of Continuing Calibration

         Prior to the analysis of samples, including MS/MSD and required
         blanks, and after tuning criteria and initial calibration criteria
         have been met, each GC/MS system must be routinely checked by
         analyzing a continuing calibration standard to ensure that the
         instrument continues to meet the instrument sensitivity and linearity
         requirements of the SOW.  The continuing calibration standard
         contains all the semivolatile target and surrogate compounds and

 -   -                              D-23/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 9
Calibration and Standardization
Continuing Calibration

         internal standards.

9.4.2    Frequency of Continuing Calibration

9.4.2.1     Each GC/MS used  for analysis must be calibrated once  every  12-hour
            time period of operation.   The  12-hour time period begins with  the
            injection of  DFTPP.

9.4.2.2     If  time  still remains  in  the 12-hour time period  after  meeting  the
            technical acceptance criteria  for the initial  calibration,  samples
            may be analyzed.   It is not necessary to analyze  a continuing
            calibration standard within this 12-hour time  period  if the
            initial  calibration standard that is the same  concentration as  the
            continuing calibration standard meets the continuing  calibration
            technical acceptance criteria.  Quantitate all sample results
            against  the initial calibration standard that  is  the  same
            concentration as the continuing calibration standard  (50 ng/2 uL).

9.4.3    Procedure for Continuing  Calibration

9.4.3.1     All standard/spiking solutions and  blanks must be allowed  to warm
            to  ambient temperature (approximately 1 hour)  before  preparation
            or  analysis.

9.4.3.2     Add a  10 uL aliquot of internal standard solution (Section
            7.2.4.6) to 1 mL aliquot  of continuing calibration standard to
            result in  40  ng of internal standard in the 2  uL  volume of
            calibration standard injected  onto  the GC/MS.  The internal
            standards  specified in. Section 7.2.4.6 should  permit  most  of the
            semivolatile  target compounds  to have relative retention times  of
             0.80  to  1.20, using the assignments of internal  standards  to
            target compounds given in Table 2.

9.4.3.3     Analyze  the continuing calibration  standard by injecting 2.0 uL of
            standard.

9.4.4     Calculations  for Continuing Calibration

9.4.4.1     Calculate  a  relative  response  factor  (RRF)  for each  semivolatile
             target and surrogate  compound using Equation  1 and the  primary
             characteristic  ions found in  Table  3  (Internal Standards)  and
             Table 4  (Target Compounds and Surrogates).   For  internal
             standards,  use  the primary ions listed  in  Table  3 unless
             interferences are present.  If interferences  prevent the use of
             the primary ion for a  given internal  standard, use the  secondary
             ion(s)  listed in Table 3.
                                    D-24/SVOA                            OLM04.2

-------
                                           Exhibit D Semivolatiles -- Section 9
                                                Calibration and Standardization
                                                         Continuing Calibration

9.4.4.2     Calculate the percent difference between the mean relative
            response factor from the most recent initial calibration and the
            continuing calibration relative response .factor for each
            semivolatile target and surrogate compound using Equation 3.

            EQ.  3.
                                         RRF  - RRF.
                        % Difference™,. =  ----- x 100
                                    ***      RRF.
                                               .
            Where,

            RRF.    =   Mean relative response factor from the most recent
                       initial calibration meeting technical acceptance
                       criteria.
            RRFC    =    Relative  response factor  from continuing  calibration
                         standard.

9.4.5    Technical Acceptance Criteria for Continuing Calibration

9.4.5.1     The continuing calibration standard must be analyzed at the  50
            ng/2 uL concentration level at the frequency described in  Section
            9.4.2, on a  GC/MS system meeting the DFTPP tuning and the  initial
            calibration  technical acceptance criteria.

9.4.5.2     The relative response factor  (RRF) for each semivolatile target
            and surrogate compound must be greater than or  equal to the
            compound's minimum acceptable relative response  factor listed in
            Table 5.

9.4.5.3     The relative response factor percent difference  for each
            semivolatile target and surrogate compound that  has a percent
            difference criteria must be within the inclusive range listed in
            Table 5.

9.4.5.4     Up to four semivolatile target compounds may fail to meet  the
            minimum RRF  or maximum percent difference criteria listed  in Table
            5, but the RRFs of those four compounds must be  greater than or
            equal to 0.010, and the percent differences must be within the
            inclusive range of ± 40.0 percent.

9.4.5.5     Excluding those ions in the solvent front, no quantitation ion may
            saturate the detector.  Consult the manufacturer's instrument
            operating manual  to determine how saturation is  indicated  for your
            instrument.

9.4.6    Corrective Action for Continuing Calibration

9.4'.6.1     If any continuing calibration technical acceptance criteria  are
          1  not met, recalibrate the GC/MS instrument according to Section
            9.3.3.  It may be necessary to clean the  ion source, change  the
            column, or take other corrective actions  to achieve the continuing
            calibration  technical acceptance criteria.

                                   D-25/SVOA                             OLM04.2

-------
Exhibit D Seraivolatiles -- Section 9
Calibration and Standardization
Continuing Calibration

9.4.6.2     Continuing calibration technical acceptance criteria MUST be met
            before any samples, including MS/MSD or required blanks, are
            analyzed.  Any samples,  including MS/MSD or required blanks,
            analyzed when continuing calibration criteria have not been met
            will require reanalysis  at no additional cost to the Agency.
                                    D-26/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles -- Section 10
                                                                       Procedure
                                                             Sample Preparation

10.0  PROCEDURE.

10.1  Sample Preparation

10.1.1   If sufficient sample amount  (less than 90% of the required amount) is
         received to perform the analyses, the Contractor shall contact SMO to
         apprise them of the problem.  SMO will contact the Region for
         instructions.  The Region will either require that no sample analyses
         be performed, or will require that a reduced volume be used for the
         sample analysis.  No other changes in the analyses will be permitted.
         The Contractor shall document the Region's decision in the SDG
         Narrative.

10.1.2   If multiphase samples (e.g., a two-phase liquid sample, oily
         sludge/sandy soil sample) are received by the Contractor, the
         Contractor shall contact SMO to apprise them of the type of sample
         received.  SMO will contact the Region.  If all phases of the sample
         are amenable to analysis, the Region may require the Contractor to do
         any of the following:

         •  Mix the sample and analyze an aliquot from the homogenized sample.

         •  Separate the phases of the sample and analyze each phase
            separately.  SMO will provide EPA sample numbers for the
            additional phases.

         •  Separate the phases and analyze one or more of the phases, but not
            all of the phases.  SMO will provide EPA sample numbers for the
            additional phases, if required.

         •  Do not analyze the sample.

10.1.2.1       If all of the phases are not amenable to analysis  (i.e.,
               outside scope), the Region may require the Contractor to do   any
               of the following:

               •  Separate the phase(s) and analyze the phase(s) that  is
                  amenable to analysis.  SMO will provide EPA sample numbers
                  for the additional phases, if required.

               •  Do not analyze  the  sample.

10:1.2.2       No other change in the analyses will be permitted.  The
               Contractor shall document the Region's decision in the  SDG
               Narrative.

10.1.3   Water Sample

10.1.3.1       Continuous liquid-liquid extraction is used to extract  the
               samples.  Separatory funnel extraction  or other manual
               extraction techniques  cannot be used.

10.1.3.2       Continuous Liquid-Liquid Extraction Without Hydrophobic
               Membrane

                                   D-27/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 10
Procedure
Sample Preparation
10.1.3.2.1         Follow manufacturer's  instructions for set-up.

10.1.3.2.2         Add methylene  chloride to  the bottom of  the  extractor and
                   fill  it  to a depth of  at least one inch  above  the bottom
                   sidearm.

10.1.3.2.3         Measure  out a  1  L sample aliquot  in a separate,  clean
                   graduated cylinder;  transfer the  aliquot to  the  continuous
                   extractor.  Measure and record the  initial pH of the sample
                   with  a pH meter or narrow  range pH paper.    Adjust  the pH to
                   2.0 with 1:1 H2SO4 and  record the final pH.   NOTE:   With
                   some  samples,  it may be necessary to place a layer  of glass
                   wool  between the methylene chloride and  the  water layer in
                   the extractor,  to prevent  precipitation  of suspended solids
                   into  the methylene chloride during extraction.

10.1.3.2.4         Using a  syringe or volumetric  pipet, add 0.5 mL  of  the
                   surrogate standard spiking solution  (Section 7.2.4.1)  into
                   the sample and mix well.

10.1.3.2.5         Rinse the graduated cylinder with 50 mL  of methylene       '
                   chloride and transfer  the  rinsate to the continuous
                   extractor.  If the sample  was  received  in a  1 L  container,
                   rinse the empty container  with 50 mL of  methylene  chloride
                   and add rinsate to the continuous extractor.

10.1.3.2.6         Add  sufficient methylene  chloride to the continuous
                   extractor to ensure proper solvent  cycling during  operation.
                   Adjust the drip rate to 5  to  15  mL/minute (recommended);
                   optimize the extraction drip  rate.  Extract  for  a   minimum of
                   18 hours.  NOTE:  When a  minimum drip  rate of 10-15 mL/min
                   is maintained throughout  the  extraction, the extraction  time
                   may be reduced to a minimum of 12 hours.  Allow  to cool  then
                   detach the distillation flask.   Proceed to Section 10.2.

10.1.3.2.7         NOTE:  Some continuous liquid-liquid extractors  are also
                   capable of concentrating the  extract within  the  extraction
                   set-up.   Follow the manufacturer's  instructions  for
                   concentration when using this  type  of  extractor.

10.1.3.3        Continuous Liquid-Liquid Extraction With  Hydrophobic  Membrane

10.1.3.3.1         Follow the manufacturer's instructions for set-up.

10.1.3.3.2         Measure out each 1 L sample aliquot in a separate,  clean
                   graduated cylinder.  If the sample  was received in a 1 L
                   container, rinse the empty container  with 50 mL of methylene
                   chloride  and add the rinsate to the continuous extractor.
                   If the sample was not  received in a 1  L container,  add 50 mL
                   of methylene chloride  to the continuous extractor.   Slowly
                   transfer  the aliquot to the continuous extractor.    Measure
                   and record the initial pH of the sample with a pH meter or
                   narrow range pH paper.  Adjust the pH to 2.0 with 1:1 H 2SO4

                                    D-28/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles --  Section 10
                                                                      Procedure
                                                             Sample Preparation

                  and record the final pH.

10 .-1.3.3.3        Using a syringe or volumetric pipet, add 0.5 mL of the
                  surrogate standard spiking solution  (Section 7.2.4.1) into
                  the sample and mix well.

10.1.3.3.4        Rinse the graduated cylinder with  50 mL of methylene
                  chloride and transfer the rinsate  to the continuous
                  extractor.

10.1.3.3.5        Add sufficient methylene chloride  to the continuous
                  extractor to ensure proper solvent cycling during operation.
                  Adjust the drip rate to 15 mL/minute  (recommended); optimize
                  the extraction drip rate.  Extract for a  minimum of  6 hours.
                   (NOTE:  Due to the smaller volume  of solvent used during  the
                  extraction process, some sample  matrices  (e.g., oily
                  samples, samples containing a high concentration of
                  surfactants) may create an emulsion which will consume the
                  solvent volume, preventing the efficient extraction of the
                  sample.  When this occurs, add additional solvent to  assure
                  efficient extraction of the sample and extend  the extraction
                  time for a minimum of  6 hours.   If the  sample  matrix
                  prevents the free  flow of solvent  through the  membrane, then
                  the non-hydrophobic membrane continuous liquid-liquid type
                  extractor must be used.)  Allow  to cool, then  detach  the
                  distillation flask.  Proceed to  Section 10.2.

10.1.3.3.6        NOTE:  Some continuous liquid-liquid extractors are also
                  capable of concentrating the extract within the extraction
                  set-up.  Follow the manufacturer's instructions for
                  concentration when using this  type of extractor.  Using the
                  hydrophobic membrane type extractor, it may not be necessary
                  to dry the extract with sodium sulfate.

10.1.3.4       NOTE:  If low surrogate recoveries  occur, assure: 1) the
               apparatus was properly assembled to prevent leaks; 2) the drip
               rate/solvent cycling was optimized; and 3) there  was proper
               cooling for condensation of solvent.

10.1.3.5       NOTE:  Alternate continuous liquid-liquid extractor types that
               meet the requirements of the SOW may  also be used.  If using
               alternate extractors or design types  follow the manufacturer's
               instruction for set-up.

10.1.4    Soil/Sediment Samples

          Decant and discard any water layer on a sediment sample.  Mix  samples
          thoroughly, especially composited samples.  Discard any foreign
          objects such as sticks, leaves,  and rocks.
                                   D-29/SVOA                             OLM04.2

-------
Exhibit D Semivolatiles --. Section 10
Procedure
Sample Preparation

10.1.4.1       pH Determination
               Transfer  50 g  of  soil/sediment to a  100 mL beaker.  Add  50 mL
               of water  and stir for  1 hour.  Determine pH of  sample with a pH
               meter while stirring.  Report pH value on appropriate data
               sheets.   If the pH of  the  soil/sediment is greater  than  11 or
               less than 5, document  this in the SDG Narrative.  Discard this
               portion of sample.  NOTE:   If limited sample  volume is
               received,  use'a smaller 1:1 ratio of grams of soil/sediment
               sample to mLs  of  water for the pH determination.  Note this in
               the SDG Narrative.
10.1.4.2        Percent  Moisture
                Immediately after weighing the  sample  for extraction,  weigh 5-
                10  g of  the soil/sediment  into  a tared crucible.   Determine the
                percent  moisture by drying overnight at 105  °C.   Allow  to  cool
                in  a desiccator before weighing.   Concentrations  of individual
                analytes will be reported  relative to  the dry weight of
                soil/sediment.
                EQ.4

                             grams of wi
                   n Ht-iiTf» = —	
                                        grams of wet sample
.  „ .         grams of wet sample - grams of dry sample   „„„
% Moisture =  —	  x 100
 10.1.4.3        Mandatory Determination of Concentration Level

                The Contractor must determine whether a soil/sediment sample
                should be analyzed by the low or medium level soil/sediment
                method.  It is the responsibility of the Contractor to analyze
                the sample at the correct level.   Three approaches may be taken
                to determine whether the low level or medium level method must
                be followed.

                •  Assume the sample is low level and analyze a 30 g sample.

                •  Use the screening method in Appendix A to determine the
                   appropriate method for analysis.

                •  Use other EPA screening procedures or  an in-house laboratory
                   screening procedure.  The procedure must be documented and
                   available for review during on-site laboratory evaluation or
                   when requested by the Technical Project Officer or
                   Administrative Project Officer.

 10.1.4.4        Low Level Soil/Sediment Samples

                The procedure described below is for the extraction of
                soil/sediment samples by sonication.  The Contractor may also
                use Automated Soxhlet Extraction  (SW-846 Method 3541
                Revision 0, September 1994) or Pressurized Fluid Extraction
                 {SW-846 Method 3545 Revision 0, December 1996) techniques for

                                   . D-30/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles --  Section 10
                                                                      Procedure
                                                             Sample Preparation

               soil/sediment samples.  The above SN-846 methods are provided
               as reference only and the laboratory supplied Standard
               Operating Procedures  (SOP) -must be-accepted-by the Agency
               before the laboratory can utilize these me'thods  (See Exhibit E
               for required IPR studies).  The requirements of this SOW must
               be met at all times  (i.e., original sample weight).  As
               applicable, follow manufacturer's instructions for the use of
               all extraction equipment.  If one of the above alternative
               extraction procedures is used, the Contractor must maintain
               documentation of the procedure utilized, and document its
               equivalence to the sonication procedure described below.  If
               the sample weight must be adjusted to utilize one of the
               alternative extraction procedures, the Contractor shall
               immediately contact SMO to inform them of the problem.  SMO
               will contact the Region for instructions.

               NOTE: All soil/sediment samples in a Case must Jbe extracted by
               the same procedure.

10.1.4.4.1        The following steps should be performed rapidly to avoid
                  loss of the more volatile extractables.   Weigh approximately
                  30 g of sample,  to the nearest 0.1 g, into a 400 mL beaker
                  and add 60 g of anhydrous powdered or granulated sodium
                  sulfate.  Mix well.  The sample should have a sandy texture
                  at this point.  Add 0.5 mL of the surrogate standard spiking
                  solution (Section 7.2.4.1)  to the sample,  then immediately
                  add 100 mL of 1:1 methylene chloride-acetone.

10.1.4.4.2        Place the bottom surface of the tip of the 3/4 inch tapered
                  disrupter horn about M inch below the surface of the
                  solvent, but above the sediment layer.

10.1.4.4.3        Sonicate for 3 minutes using a 3/4 inch disrupter horn at
                  full power (output control knob at 10) with pulse on and
                  percent duty cycle knob set•at 50.0 percent.  Do  not use a
                  microtip.   NOTE:  These settings refer to the Model W-385.
                  When using a sonicator other than Model W-385, refer to the ,
                  instructions provided by the manufacturer for appropriate
                  output settings.  Decant and filter extracts through Whatman
                  #41 filter paper using vacuum filtration or centrifuge and
                  'decant extraction solvent.

10.1.4.4.4        Repeat the extraction two more times with 2 additional 100
                  mL portions of 1:1 methylene chloride-acetone.  Before each
                  extraction,  make certain that the sodium sulfate is free-
                  flowing and not a consolidated mass.  As required,  break up
                  large lumps with a clean spatula, or, very carefully,  with
                  the tip of the unenergized probe.  Decant the extraction
                  solvent after each sonication.  On the final sonication,
                  pour the entire sample into the Buchner funnel and rinse
                  with 1:1 methylene chloride-acetone.

10.1.4.4.5        If the sample is to be screened following the low level
                  preparation method prior to GPC, refer to Appendix A,

                                   D-31/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 10
Procedure
Sample Preparation

                  "Screening of Seraivolatile Organic Extracts."

10.1.4.4.6        After screening, transfer the remainder of the 1 mL back to
                  the total extract.  CAUTION:  To minimize sample loss,
                  autosamplers which pre-flush samples through the syringe,
                  should not be used.

10.1.4.5       Medium Level  Soil/Sediment  Samples

               The procedure described  below  is for  the  extraction of
               soil/sediment samples by sonication.   The Contractor  may also
               use Automated Soxhlet Extraction  (SH-846  Method 3541
               Revision 0,  September 1994)  or Pressurized Fluid  Extraction
                (SN-846  Method 3545 Revision 0, December  1996)  techniques for
               soil/sediment samples.   The above SN-846  methods  are  provided
               as reference only and the laboratory  supplied  Standard
               Operating Procedures  (SOP)  must be  accepted by the Agency
               before  the laboratory can utilize these methods (See  Exhibit E
               for required IPR studies).   The requirements of this  SOW must
               be met  at all times (i.e.,  original sample weight).   As
               applicable,  follow manufacturer's instructions for  the use of
               all extraction equipment.   If  one of  the  above alternative
               extraction procedures is used,  the  Contractor  must  maintain
               documentation of the procedure utilized,  and document its
               equivalence to the sonication  procedure described below.   If
               the sample weight must be adjusted  to utilize  one of  the
               alternative extraction procedures,  the Contractor shall
               immediately contact SMO  to inform  them of the  problem.   SMO
               will  contact the Region  for instructions.

               NOTE: All soils/sediment samples in a Ca.se must be  extracted by
                the  same procedure.

10.1.4.5.1         Transfer approximately 1 g  (record weight to the nearest
                   0.1  g) of sample to  a 20 mL vial.   Wipe the mouth of  the
                  vial with a tissue to remove any sample material.   Record
                   the  exact weight of  sample  taken.   Cap the  vial  before
                  proceeding with the  next sample  to avoid any cross-
                   contamination.

10.1.4.5.2        Add 2.0 g of anhydrous powdered or granulated sodium sulfate
                   to the sample in the  20 mL  vial  and mix well.

10.1.4.5.3         Surrogates are added to all samples, spikes, and blanks.
                   Add 0.5 mL of surrogate spiking solution (Section 7.2.4.1)
                   to the sample mixture.

10.1.4.5.4         Immediately add 9.5 mL of methylene chloride to the sample
                   and disrupt the sample with the 1/8 inch tapered Microtip
                   ultrasonic probe for 2 minutes  at output control setting 5,
                   in continuous mode (if using a sonicator other than Models
                   W-375 or W-385, contact the instrument manufacturer for
                   appropriate output settings).   Before extraction,  make
                   certain that the sodium sulfate is free-flowing and not a

  .    .            .                  D-32/SVOA                             OLM04.2

-------
                                          Exhibit D Semivolatilss -- Section 10
                                                                       Procedure
                                                           Sample Concentration

                  consolidated mass.   Decant and filter extract through
                  Whatman #41 filter paper using vacuum filtration  or
                  centrifuge-and decant extraction  solvent.-

10.2  Concentrating the Extract

      NOTE:  Low level soil/sediment samples prepared by the procedure
      described in Section 10.1.4.4 will result  in extracts containing a
      mixture of acetone and methylene chloride.  Because all soil/sediment
      sample extracts MUST be subjected to GPC cleanup prior to analysis, the
      majority of the acetone must be removed from the extract, otherwise, it
      will have adverse effects on the GPC column.  To remove the acetone from
      the soil/sediment sample extract, follow the steps in Section 10.2.1
      then concentrate to 1 mL using the nitrogen evaporation technique in
      Section 10.2.2.2.

10.2.1   Concentration by K-D

10.2.1.1       Assemble a Kuderna-Danish  (K-D) apparatus by attaching  a  10 mL
               concentrator tube to a  500 mL evaporative flask.  Other
               concentration devices or techniques  may be used  in place  of the
               K-D, if equivalency is  demonstrated  for all the  semivolatile
               target compounds listed in Exhibit C.

10.2.1.2       For water samples, transfer the extract to a Kuderna-Danish  (K-
               D) concentrator by pouring the extract through a drying column
               containing about 10 cm  of anhydrous  granular sodium  sulfate.

10.2.1.2.1        For soil/sediment samples, directly transfer  the  extract to
                  the K-D concentrator.

10.2.1.2.2        Rinse the Erlenmeyer flasks  (for  both water and
                  soil/sediment samples) and the column  (for water  samples)
                  with 20-30 mL of methylene chloride to complete the
                  quantitative transfer.

10.2.1.3  •     Add one or two clean boiling  chips to the evaporative flask and
               attach a three-ball Snyder column.   Pre-wet the  Snyder  column
               by adding about 1 mL methylene chloride to the top of the
               column.  Place the K-D  apparatus  in  a hot water  bath (60   °C to
               70 °C recommended) so that.the concentrator tube is  partially
               immersed in the hot water and the entire lower rounded  surface
               of the flask is bathed  with hot vapor.  Adjust the vertical
               position of the apparatus and the water temperature  as  required
               to complete the concentration in  10  to 15 Minutes.   At  the
               proper rate of distillation,  the  balls of the column will
               chatter actively, but the chambers will not flood with
               condensed solvent.  When the  apparent volume of  liquid  reaches
               1 mL, remove the K-D apparatus from  the water bath and  allow it
               to drain and cool for at least 10 minutes.  DO NOT ALLOW  THE
               EVAPORATOR TO GO DRY.   Remove the Snyder column  and  rinse  the
               flask and its lower joint into the concentrator  tube with 1 or
               2 mL of methylene chloride.   A 5-mL  syringe is recommended for
               this operation.

                                   b-33/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles. -- Section 10
Procedure
Sample Concentration
10.2.1.4        For water  samples  which do not  require GPC  cleanup, proceed to
                final  concentration of extract  (Section  10.2.2).   Oily water
                samples  require GPC.

10.2.1.5        For water  samples  which require GPC,  adjust the volume of  the
                extract  to 10.0 mL with methylene  chloride  and proceed with GPC
                cleanup  (Section 10.3).

10.2.1.6        For soil/sediment  samples, adjust  the volume of the extract to
                10.0 mL  with methylene chloride, and proceed with GPC cleanup
                (Section 10.3).

10.2.1.7        For water  samples  or soil/sediment samples  which  have undergone
                GPC, proceed to final concentration of  extract (Section
                10.2.2) .

10.2.2    Final  Concentration of Extract

          Two  different  concentration techniques are permitted to obtain the
          final  extract  volume, Micro Snyder column and Nitrogen  Evaporation
          Techniques:

10.2.2.1        Micro  Snyder Column Technique

                Add another one or two  clean boiling chips  to the concentrator
                tube  and attach a two-ball micro Snyder column.   Pre-wet the
                Snyder column by adding about 0.5  mL of methylene chloride to
                the top of the column.   Place the  K-D apparatus  on a  hot water
                bath  (60  °C to 70  "C recommended)  so that the concentrator tube
                is partially immersed in the hot water.  Adjust  the vertical
                position of the apparatus and the  water temperature as  required
                to complete the concentration in 5 to 10 minutes.  At the
                proper rate of distillation, the balls of the column  will
                chatter actively,  but the chambers will not flood with
                condensed solvent.  When the apparent volume of  liquid reaches
                about 0.5 mL  (0.4 mL for soil/sediment samples or water samples
                that have undergone GPC), remove the K-D apparatus from the
                water bath and allow it to drain for at least 10 minutes while
                cooling.  Remove the Snyder  column and rinse the evaporative
                flask and its  lower joint into the concentrator tube  with 0.2
                mL (O.I mL for soil/sediment samples and water samples that
                have undergone GPC) of  methylene chloride.- Adjust the final
                volume to 1.0 mL  (0.5 mL for soil/sediment samples and water
                samples that have undergone  GPC) with methylene chloride.
                Transfer  the extract to the  Teflon-sealed screw-cap bottle,
                label the bottle, and store at  4°C (±  2°C).

 10.2.2.2        Nitrogen  Evaporation Technique  (Taken from ASTM Method D3086)

                The following  method may be used for final concentration of the
                semivolatile extract instead of the procedure in Section
                10.2.2.1,  Place  the concentrator tube in a warm water bath
                (30 °C to 35  °C recommended) and evaporate the solvent volume

                                    D-34/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles -- Section 10
                                                                      Procedure
                                                           Sample Cleanup.(GPC)

               to just below 1 mL  (below 0.5 mL  for soil/sediment samples  and
               water samples that have undergone GPC) by blowing a gentle
               stream of clean, dry nitrogen  (filtered through a column of
               activated carbon) above the extract.  CAUTION:  Gas lines from
               the gas source to the evaporation apparatus must be stainless
               steal, copper, or Teflon tubing.  New plastic tubing must not
               be used between the carbon trap and the sample, since it may
               introduce interferences.  The internal wall of the concentrator
               tube must be rinsed down several  times with methylene chloride
               during the operation.  During evaporation, the tube solvent
               level must be kept below the water level of the bath.  The
               extract must never be allowed to  become dry.
10.2.2.3       Final Extract Volumes

10.2.2.3.1        Water
                  For water  samples that did not undergo GPC, the extract must
                  be brought to a final volume of  1.0 mL with methylene
                  chloride.  For water samples that underwent GPC, the extract
                  must be brought to a final volume of  0.5 mL with methylene
                  chloride  (concentrating the extract to 0.5 mL will result in
                  no loss of sensitivity despite the volume of extract  (5 mL)
                  not recovered after GPC).

10.2.2.3.2        Soil/Sediment

                  Adjust the final volume to 0.5 mL with methylene chloride
                   (concentrating the extract to 0.5 mL  will result in no loss
                  of sensitivity despite the volume of  extract  (5 mL) not
                  recovered  after GPC).

10.2.2.3.3        Transfer the extract to a Teflon-sealed screw-cap bottle,
                  label the  bottle, and store at  4°C (± 2"C).

10.3  Sample Cleanup by GPC

10.3.1   Introduction

10.3.1.1       Gel Permeation Chromatography (GPC) is a size exclusion cleanup
               procedure using organic solvents and hydrophobic gels in the
               separation of natural macromolecules.  The packing gel is
               porous and is characterized by the  range or uniformity
                (exclusion range) of that pore size.  In the choice of gels,
               the exclusion range must be larger  than  the molecular size of
               the molecules to be separated.
10.3.1.2       GPC must  be  performed  for  all  soil/sediment  extracts.  GPC   must
               be performed for water extracts that  contain higher molecular
               weight  contaminants that interfere with the  analysis of the
               target  analytes.
                                   D-35/SVOA                             OLM04.2

-------
Exhibit D Semivolatiles -- Section 10
Procedure
Sample Cleanup  (GPC)

10.3.2   GPC Column Preparation

         The instructions  listed below  for GPC column preparation are for Bio
         Beads.  Alternative  column packings may be used  if  1)  the column
         packings have  equivalent or better performance than the Bio Beads and
         meet the technical acceptance  criteria for GPC calibration and   GPC
         continuing calibration verification ,2)  the column packings do not
         introduce  contaminants/artifacts into the sample which interfere with
         the analysis of  the  semivolatile compounds.  Follow the
         manufacturer's instructions for preparation of the  GPC column
         packing.

10.3.2.1        Weigh  out  70  g of  Bio Beads  (SX-3).  Transfer them  to  a quart
                bottle with a Teflon-lined cap  or  a  500 mL separatory  funnel
                with a large  bore  stopcock,  and add  approximately 300  mL  of
                methylene  chloride.   Swirl the  container  to ensure  the wetting
                of  all beads.   Allow the beads  to  swell for a minimum  of  2
                hours.   Maintain enough  solvent to cover  the beads  sufficiently
                at  all times.   If  a  guard column  is  to be  used,  repeat the
                above  with 5  g of  Bio Beads  in  a  125 mL bottle or a beaker
                using  25 mL of the methylene chloride.

10.3.2.2        Turn the column upside down  from  its normal position and  remove
                the inlet bed support plunger (the inlet  plunger is longer  than
                the outlet plunger).   Position  and tighten the outlet  bed
                support  plunger as near  the  end as possible,  but no closer  than
                5 cm (measured from the  gel  packing  to  the collar).

10.3.2.3        Raise  the end of the outlet  tube  to  keep  the solvent  in  the GPC
                column,  or close the column  outlet stopcock.   Place a  small
                amount of solvent in the column to minimize the formation of
                air bubbles at the base  of poured column  packing.

10.3.2.4        Swirl  the bead/solvent  slurry to  get a  homogeneous  mixture  and,
                if the wetting was done  in  a quart bottle, quickly transfer it
                to a 500 mL separatory funnel with a large bore stopcock.
                Drain the excess methylene  chloride  directly into the  waste
                beaker,  and start draining  the  slurry into the column  by
                placing the separatory funnel tip against the column wall.
                This will help to minimize  bubble formation.  Swirl
                occasionally to keep the slurry homogeneous.  Drain enough to
                fill the column.  Place the  tubing from .the column outlet into
                a waste beaker below the column,  open the stopcock (if
                attached), and allow the excess solvent to drain.   Raise the
                tube to stop the flow,  and close the stopcock when the top of
                the gel begins to look dry.   Add additional methylene  chloride
                to just rewet the gel.

 10.3.2.5   ..    Wipe any remaining beads and solvent from the inner walls of
                the top of the column with a laboratory tissue.  Loosen the
                seal slightly on the other plunger assembly  (long plunger)  and
                insert  it  into the column.   Make the seal just tight enough so
                that any beads on the glass surface will be pushed forward, but
                loose enough so that the plunger can be pushed forward.

                                    D-36/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles -- Section 10
                                                                       Procedure
                                                           Sample Cleanup  (GPC)

               CAUTION:  Do not tighten  the  seal  if beads  are between  the  seal
               and  the glass  surface because, this can  damage the seal  and
               cause--leakage.      ._.....       ...     _  .

10.3.2.6       Compress  the column  as much as possible without  applying
               excessive force.  Loosen  the  seal  and gradually  pull  out the
               plunger.  Rinse and  wipe  off  the plunger.   Slurry any remaining
               beads  and transfer them into  the column.  Repeat the  step in
               Section 10.3.2.5 and reinsert the  plunger.  If the plunger
               cannot be inserted and pushed in without allowing beads to
               escape around  the seal, continue compression of  the beads
               without tightening the seal,  and loosen and remove the  plunger
               as described.  Repeat this procedure until  the plunger  is
               inserted  successfully.

10.3.2.7       Push the  plunger until it meets the gel, then compress  the
               column bed  about four centimeters.

10.3.2.8       Pack the  optional 5  cm column with approximately 5 g  of pre-
               swelled beads  (different  guard columns  may  require different
               amounts).   Connect, the guard  column to  the  inlet of the
               analytical  column.

10.3.2.9       Connect the column inlet  to  the solvent reservoir  (reservoir
               should be placed higher than  the top of the column) and place
               the  column  outlet tube in a waste  container.  Placing a
               restrictor  in  the outlet  tube will force air out of the column
               more quickly.  A restrictor  can be made from a piece  of
               capillary stainless  steel tubing of 1/16" OD x O.010" ID x  2".
               Pump methylene chloride through the column  at a  rate  of 5
               mL/min for  one hour.

10.3.2.10      After washing  the column  for  at least one hour,  connect the
               column outlet  tube,  without  the restrictor, to the inlet side
               of the UV detector.  Connect  the system outlet to the outlet
               side of the UV detector.  A  restrictor  (same size as  the one in
               Section 10.3.2.9) in the  outlet tube from the UV detector will
               prevent bubble- formation  which causes a noisy UV baseline.

10.3.2.11      The  restrictor will  not affect flow rate.   After pumping
               methylene chloride through the column for an additional 1-2
               hours, adjust  the inlet bed  support plunger until approximately
               6-10 psi  backpressure is  achieved. Push the plunger  in to
               increase  pressure or slowly pull outward to reduce pressure.

10.3.2.12    '  When the  GPC column  is not to be used for several days,  connect
               the  column  outlet line to the column inlet  to prevent column
               drying and/or  channeling. If channeling occurs, the"gel must
               be removed  from the  column,  re-swelled', and re-poured as
               described above.  If drying  occurs, methylene chloride  should
               be pumped through the column  until the  observed  column  pressure
               is constant and the  column appears wet.  Always  recalibrate
               after column drying  has occurred to verify  retention  volumes
               have not  changed.

                                    D-37/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 10
Procedure
Sample Cleanup (GPC)

10.3.3   Calibration of GPC

10.3.3.1       Summary of  GPC Calibration and GPC Continuing Calibration
               Veri ficat ion  (CCV)

               The  GPC calibration procedure  and GPC CCV procedure are  based
               on monitoring the elution  of standards with  a UV  detector
               connected  to  the GPC  column.

10.3.3.2       Frequency  of  GPC Calibration and GPC CCV

               Each GPC system must  be. calibrated upon  award of  a  contract,
               when the column is changed, when channeling  occurs,  and once
               every seven days  CGPC Continuing. Calibration Verification,)  when
               samples, including matrix  spikes, matrix spike  duplicates,  and
               blanks,  are cleaned up using GPC.

10.3.3.3       Procedure  for GPC Calibration  and GPC CCV

               Follow  the manufacturer's  instructions  for operating the GPC
               system.   Changes  in pressure,  solvent flow rate,  and
                temperature conditions can affect analyte retention times and
               must be monitored.

10.3.3.3.1         Using a 10 mL syringe,  load the calibration solution
                    (Section 7.2.4.3)  onto  the  GPC.   Establish appropriate
                   collect and dump  time periods to ensure collection of all
                   target  analytes.   Initiate  column eluate collection just
                   before elution of bis  (2-ethylhexyl)  phthalate and after the
                   elution of corn oil.  Stop  eluate collection shortly after
                   the  elution of perylene.  Collection should be stopped
                   before sulfur elutes.  Use  a "wash" time of 10 minutes after
                   the elution of sulfur.   Each laboratory is required to
                  .establish its specific time sequences.

10.3.3.3.2         Reinject the calibration solution after appropriate collect
                   and dump cycles have been set, and the solvent flow and
                   column pressure have been established.

10.3.3.3.3         Measure and record the volume of collected GPC eluate in a
                   graduated cylinder.  The volume of GPC eluate collected for
                   each sample extract processed may be used to indicate
                   problems with the system during sample processing.

10.3.3.3.4         Analyze a GPC blank of methylene chloride  after each GPC
                    calibration or each GPC CCV .  Concentrate  the methylene
                    chloride that passes through the system during the collect
                    cycle using a Kuderna-Danish  (K-D) evaporator.  Add internal
                    standards at the appropriate concentration and analyze the
                    concentrate by GC/MS.
                                    D-38/SVOA                            OLM04.2

-------
                                            Exhibit D Semivolatiles -- Section 10
                                                                         Procedure
                                                              Sample Cleanup  (GPC)

|  10.3.4   Technical Acceptance Criteria for GPC Calibration   and GPC CCV

|  10.3.4.1        The GPC  system must  be calibrated and verified  at the  frequency
                  described in Section 10.3.3.2.  The UV trace must meet  the
                  following requirements:

                  •   Peaks must be observed and should  be  symmetrical  for all
                     compounds in  the  calibration solution.

                  •   Corn  oil and  the  phthalate peaks must exhibit > 85.0 percent
                     resolution.

                  •   The phthalate and methoxychlor peaks  must exhibit >  85.0
                     percent resolution.

                  •   Methoxychlor  and  perylene  peaks must  exhibit  > 85.0  percent
                     resolution.

                  •   Perylene and  sulfur peaks  must not be saturated and  must
                     exhibit > 90.0 percent baseline resolution.

  10.3.4.2        The solvent flow rate and column  pressure must be within the
|                  manufacturer's specified  ranges .

  10.3.4.3        If  the retention time shift  is >  5.0  percent between
                  calibrations take corrective  action.  Excessive  retention time
                  shifts are caused by the  following:

                  •   Poor  laboratory temperature control or system leaks.

                  •   An unstabilized column that requires  pumping  methylene
                     chloride through  it for several more  hours or overnight.

                  •   Excessive laboratory temperatures  causing outgassing of  the
                     methylene chloride.

|  10.3.4.4        A  copy of the two most recent UV traces of  the  calibration
                  solution must be submitted with the data for the associated
                  samples.

|  10.3.4.5        The analyte concentrations in the GPC blank  must be less than
                  the CRQL for all target compounds in  Exhibit C  (Semivolatiles),
|                  except phthalate esters,  which must   be  less than 5  times the
                  CRQL.

|  10.3.5   Corrective  Action for  GPC  Calibration   and  GPC CCV

  10.3.5.1        If  the requirements  in Section 10.3.4 cannot be  met,  the column
                  may be cleaned by processing  several  5 mL volumes of
                  butylchloride through the system. Butylchloride removes the
                  discoloration and particles  that  may  have precipitated  out  of
                  the methylene chloride extracts.   If  a guard column  is  being
                  used, replace it with a new one.   This may correct the  problem.
                  If  column maintenance does not restore the performance  of the

                                      D-39/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 10
Procedure
Extract Cleanup  (GPC)

                column,  the  column must be  repacked with new packing  and
                recalibrated.   It may be necessary to  obtain a  new  lot of Bio
                Beads if the column  fails all  criteria.

10.3.5.2        If  the GPC blank exceeds the requirements  in  Section 10.3.4.5,
                pump  additional methylene chloride through the  system for 1-2
                hours.  Analyze another GPC blank to ensure the system is
                sufficiently clean.   Repeat the  methylene  chloride  pumping  if
                necessary.

10.4   Sample Extract Cleanup by GPC

10.4.1    It  is  very  important  to have constant  laboratory temperatures during
          an  entire GPC run,  which could be 24 hours or more.   If temperatures
          are not constant,  retention times will shift, and the dump  and
          collect times determined by the calibration  standard  no longer will
          be  appropriate.  The  ideal  laboratory  temperature to  prevent
          outgassing  of the  methylene chloride is 22   °C.

10.4.2    In  order  to prevent overloading of the GPC column, highly viscous
          sample extracts must  be diluted prior  to cleanup. Any sample extract
          with  a viscosity greater than that of  a 1:1  glycerol:water  solution
          must  be diluted and loaded into several loops.   Similarly,  extracts
          containing  more than  40 mg/mL of  non-volatile residue must  be diluted
          and loaded  into several loops.  The  non-volatile residue  may be
          determined  by evaporating' a 100 uL aliquot of the extract to dryness
          in  a  tared  aluminum weighing pan, or other suitable container.
          Systems using automated injection devices to load the sample on the
          column must be carefully monitored to  assure that the required amount
          is  being  injected  on  the column.  Viscous extracts or extracts
          containing  a large amount  of non-volatile residue will cause problems
          with  an automated  injection system's ability to  inject the  proper
          amount of sample extract on a  column.   After the sample  extract has
          been  processed, the remaining  sample extract in  the  injection vial
          must  be checked before proceeding with extract  cleanup to assure  that
          the proper  amount  was injected on the  column.   If the proper amount
          of  extract  was not injected, the  sample must be  re-prepared at no
          additional  cost to the Agency,  and  the sample extract must  either be
          diluted and loaded into  several  loops, or  the sample  extract must be
          injected manually.

          NOTE:  When multiple  loops/runs are  necessary for an  individual
          sample, be  sure to combine all  of the  sample eluates  collected  from
          each run.

 10.4.3   Frequency of GPC Sample  Cleanup

          GPC cleanup must be performed once  for each soil/sediment extract and
          for water extracts that  contain high molecular weight contaminants
          that interfere with the  analysis of  the target  analytes.   In
          addition, GPC must be performed for all associated blanks,  matrix
          spikes, and matrix spike duplicates.  If cleanup procedure  is
          inadequate, contact SMO.
                                    D-40/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles -- Section 10
                                                                       Procedure
                                                          Extract Cleanup  (GPC)

10.4.4   Procedure for GPC Sample Cleanup

10.4.4.1— -  --Particles greater-than 5-microns may scratch-the  valve, which
               may result in a system leak and cross contamination  of  sample
               extracts in the sample loops.  To  avoid  such problems,  filter
               the extract through a 5 micron filter disc by attaching a
               syringe filter assembly containing the filter disc to a 10  mL
               syringe.  Draw the sample extract  through the filter assembly
               and into the 10 mL syringe.  Disconnect  the filter assembly
               before transferring the sample extract into a small  glass
               container, e.g.,  a 15 mL culture tube with a Teflon-lined screw
               cap.  Alternatively, draw the extract into the  syringe  without
               the filter assembly.  Attach the filter  assembly  and force  the
               extract through the filter and into the  glass container.  Draw
               a minimum of 8 mL of extract' into  a 10 mL syringe.

               NOTE:  Some GPC instrument manufacturers recommend using a
               smaller micron size filter.  Follow the  manufacturer's
               recommended operating instructions.

10.4.4.2       Introduction of particulates or glass wool into the  GPC
               switching valves  may require factory repair of  the apparatus.

10.4.4.3       Follow the manufacturer's instructions for operation of the GPC
               system being utilized.   A 2 mL injection loop may be used in
               place of a 5 mL injection loop.  If a 2  mL injection loop is
               used, concentrate the sample extract to  4 mL instead of 10  mL
               and  inject 2 mL  instead of 5 mL.

10.4.4.4       If the sample is  difficult to load, part of the system  may  be
               blocked.  Take appropriate corrective action following
               manufacturer's recommendations.  The problem must be resolved
               prior to loading  sample extracts.

10.4.4.5       After loading each sample loop, wash the loading  port with
               methylene chloride to minimize cross-contamination.  Inject
               approximately 10  mL of methylene chloride to rinse the  common
               tubes.

10.4.4.6       After loading all the sample loops, process each  sample using
               the collect and dump cycle  time established in  Section
               10.3.3.3.1.

10.4.4.7       Collect each sample in a 250 mL Erlenmeyer flask, covered with
               aluminum foil to  reduce solvent evaporation, or directly into a
               Kuderna-Danish  (K-D) evaporator.   Monitor sample  volumes
               collected.  Changes in sample volumes collected may  indicate
               one or more of the following problems.

               •  Change in solvent flow rate, caused by channeling in the
                  column or changes in column pressure.

               •  Increase in column operating pressure due to the  absorption
                  of particles or gel fines onto  either the guard column or

                                   D-41/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 10
Procedure
Final Concentration/Sample Analysis by GC/MS

                   the  analytical  column gel,  if  a  guard  column  is not used.

                •  Leaks  in" the  system  or  significant  variances  in room
                   temperature.

10.4.4.8        Any samples that were loaded into two  or  more  loops  must be
                recombined before  proceeding with concentration.

10.5  Final Concentration

      Concentrate  the  extract as  per Section  10.2.2.  Final volumes should be
      brought  to the volumes stated in Section 10.2.2.3.

10.6  Sample Analysis  by  GC/MS

10.6.1    Sample extracts  shall  be analyzed only  after the GC/MS  system has met
          the  instrument performance check,  initial calibration,  and continuing
          calibration  requirements.  The same  instrument  conditions  must be
          employed  for  the analysis of  samples as were used  for  calibration.

10.6.2    Internal  standard solution is added to  each  sample extract.  Add 10
          uL of internal  standard  solution to each  accurately  measured 1.0 mL
          of water  sample  extract. For soil/sediment  samples  and water samples
          subjected to  GPC, add  5  uL of internal  standard solution to each
          accurately measured 0.5  mL of sample extract.   This  will result  in a
          concentration of 20 ng/uL of  each internal standard.

          NOTE:  An alternate amount of internal  standard solution may be
          added; however,  the internal  standards  must  be  added to maintain the
          required  20  ng/uL of each internal standard  in  the sample  extract.

10.6.3    Prior to  the  addition  of internal standards, make  any extract
          dilution  indicated by  characterization  step. Add  internal standards
          after dilution.   Internal standards must  be  added  to maintain the
          required  20  ng/uL of each internal standard  in  the extract volume.

10.6.4    Inject 2  uL  of the sample extract into  the GC/MS.  This 2  uL volume
          must contain 40 ng of  each  internal standard.

10.6.5    Sample Dilutions

10.6.5.1     If the on-column concentration of any   target  compound  in  any sample
             exceeds the initial calibration range, that  sample extract must  be
             diluted,  the internal standard concentration must be readjusted,  and
             sample extract must be reanalyzed.   Guidance in performing dilution
             and exceptions to this requirement  are given below.

10.6.5.2        Use the results of the original analysis  to determine  the
                approximate dilution factor required to get the largest analyte
                peak within the initial calibration range.
                                    D-42/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles -- Section 10
                                                                       Procedure
                                                       Sample Analysis by GC/MS

10.6.5.3       The dilution  factor  chosen  should keep  the  response  of  the
               largest peak  for a target compound  in the upper  half of the
               calibration range of the instrument-.-  -     _

10.6.5.4       The maximum dilution factor permitted for low level  soils is
               30.   If a  low level  soil sample  requires a  dilution  factor
               greater than  30, to  bring target compounds  within the
               calibration range, then the medium  level method  shall be
               utilized.

10.6.5.5       Do not submit data for more than two analyses, i.e.,   from  the
               original sample extract and one  dilution, or  if  the
               semivolatile  screening procedure was employed, from  the most
               concentrated  dilution analyzed and  one  further dilution.
                                   D-43/SVOA                             OLM04.2

-------
Exhibit D Semivolatiles -- Section 11
Data Analysis and Calculations
Qualitative Identification

11.0  DATA ANALYSIS AND CALCULATIONS

11.1  Qualitative Identification

11.1.1    Identification of Target Compounds

11.1.1.1        The  compounds  listed  in  the Target  Compound List  (TCL),  Exhibit
                C (Semivolatile),  shall  be identified by an analyst  competent
                in the interpretation of mass  spectra (see  Exhibit A,  Section
                4.3.1)  by  comparison  of  the sample  mass  spectrum  to  the  mass
                spectrum of  the standard of the suspected compound.   Two
                criteria must  be satisfied to  verify the identifications.

                • Elution of  the sample analyte within  GC  relative  retention
                  time unit window  established from the 12-hour  calibration
                  standard.

                • Correspondence of  the sample analyte  and calibration
                  standard component mass  spectra.

11.1.1.2        For  establishing correspondence of  the GC relative  retention
                time (RRT),  the sample component RRT must compare within ±0.06
                RRT  units  of the RRT of the  standard component.   For samples
                analyzed  during the  same 12-hour time period as the  initial
                calibration standards, compare the  analyte  retention times to
                those from the 50 ng calibration standard.   For reference, the
                standard  must  be run on the  same shift  as the sample.  If co-
                elution of interfering components prohibits accurate assignment
                of the sample  component RRT  from the total  ion chromatogram,
                the  RRT should be assigned by using extracted ion current
                profiles  for ions unique to  the component of interest.

11.1.1.3        For  comparison of standard and sample component mass spectra,
                mass spectra obtained from a calibration standard on the
                Contractor's GC/MS meeting the daily instrument performance
                requirements for DFTPP are required.  Once obtained, these
                standard spectra may be used for identification purposes only
                if the contractor's GC/MS meets the DFTPP daily instrument
                performance requirements.

11.1.1.4        The requirement for qualitative verification by comparison of
                mass spectra are as follows:

                •  All ions present in the standard mass spectrum at a relative
                   intensity greater  than 10.0 percent  (most abundant ion in
                   the spectrum equals 100.0 percent) must be present in the
                   sample spectrum.

                •  The relative intensities of ions specified in the paragraph
                   above must  agree within ±20.0 percent between the standard
                   and sample  spectrum,  (Example:  For  an  ion with  an abundance
                   of 50.0 percent in the standard spectrum, the corresponding
                   sample ion  abundance  must be between 30.0 and 70.0 percent).
                                    D-44/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles -- Section 11
                                                 Data Analysis and Calculations
                                                     Qualitative Identification

               •  Ions greater than 10.0 percent  in the sample spectrum, but
                  not present in the standard spectrum, must be  considered  and
                  accounted for by the analyst making the comparison.   In Task
                  II  (Exhibit A), the verification process  should  favor false
                  positives.  All compounds meeting the identification
                  criteria must be reported with  their spectra.  When target
                  compounds are below contract required quantitation limits
                  (CRQL) but the spectrum meets the identification criteria,
                  report the concentration with a "J".  For example, if the
                  CRQL is 10 ug/L and concentration of 3 ug/L is calculated,
                  report as "3J".

11.1.1.5       If a compound cannot be verified by all of the criteria  in
               Section 11.1.1  (11.1.1.1. through 11.1.1.4),  but in  the
               technical judgement of the mass spectra interpretation
               specialist, the identification is  correct, then the contractor
               shall report the identification and proceed  with  quantitation.

11.1.2   Identification of Non-Target Compounds

11.1.2.1       A 'library search shall be executed for non-target sample
               components for the purpose of tentative identification.   For
               this purpose, the NIST/EPA/NIH  (May 1992 release  or later)
               and/or Wiley (1991 release or later), or equivalent mass
               spectral library shall be used.

11.1.2.2       Up to 30 organic compounds of greatest apparent concentration
               not listed in Exhibit C for the volatile and semivolatile
               fraction, excluding the surrogate  and internal standard
               compounds, shall be identified tentatively via a  forward search
               of the NIST/EPA/NIH (May 1992 release or later) and/or Wiley
                (1991 release or later), or equivalent mass  spectral library.  ..
               The- following are not  to  be  reported :  1) Substances with
               responses less than 10 percent of  the internal standard  (as
               determined by inspection of the peak areas or heights);  2)
               substances which elute earlier than 30 seconds before the first
               semivolatile compound listed in Exhibit C  (Semivolatiles), or
               three minutes after the last semivolatile compound  listed in
               Exhibit C has eluted; and 3) volatile compounds listed in
               Exhibit C.  Only after visual comparison of  sample  spectra to
               spectra resulting from the library search(es) will  the mass
               spectral interpretation specialist assign a  tentative
               identification.  NOTE:  Computer generated library  search
               routines must not use normalization routines that would
               misrepresent the library or unknown spectra  when  compared to
               each other.

11.1.2.3       Up to 20 peaks of greatest apparent concentration (as
               determined by inspection of peak areas or heights)   that are
               suspected to be straight-chain, branched, or cyclic alkanes,
               alone or part of an  al.ka.ne series shall be library searched.
               Documentation for the tentative identification must be
               supplied.  Alkanes-are not counted as part of the 30 organic
               compounds described in Section 11.1.2.2.

                                   D-45/SVOA                             OLM04.2

-------
Exhibit D Semivolatiles -- Section 11
Data Analysis and Calculations
Qualitative Identification

11.1.2.4        Peaks  that are  suspected  to be  aldol-condensation reaction
                products  (i.e.,  4-methyl-4-hydroxy-2-pentanone and 4-methyl-3-
                pentene-2-one)  shall  be searched,  reported, and counted  as part
                of  the 30 most  intense non-target  semivolatile compounds, and
                qualified with  an  "A" flag on. Form I TIC.

11.1.2.5        Guidelines for  Making Tentative Identifications

11.1.2.5.1         Major  ions in the  reference  spectrum (ions greater than  10
                   percent of the  most abundant ion, should be present in the
                   sample  spectrum.

11.1.2.5.2         The relative intensities of  the major ions should agree
                   within  ±20  percent  (Example:  For  an ion with  an abundance
                   of  50  percent in the  standard spectrum,  the corresponding
                   sample ion abundance must  be between 30  and 70  percent).

11.1.2.5.3         Molecular ions  present in  the reference  spectrum  should  be
                   present in the  sample  spectrum..

11.1.2.5.4         Ions present in the  sample spectrum, but not  in the
                   reference spectrum,  should be reviewed for possible
                   background contamination or presence of  co-eluting
                   compounds.

11.1.2:5.5         Ions present in the  reference spectrum,  but not in the
                   sample spectrum,  should be reviewed for  possible  subtraction
                   from the_ sample spectrum because of background  contamination
                   or co-eluting compounds.   NOTE:  Data system  library
                   reduction programs .sometimes can create  these discrepancies.

11.1.2.5.6         If, after careful review and in the technical judgement of
                   the mass spectral interpretation specialist,  no valid
                   tentative identification can be made, the  compound should be
                   reported as follows:

                   9  If the library search produces a match at  or above 85%,
                      report that compound.

                   •  If the library search produces more than one compound at
                      or above 85%, report the  first compound (highest).

                   •  If the library search produces no matches at or above
                      85%, the compound  should  be reported as unknown.  The
                      mass spectral specialist  should give additional
                      classification of  the unknown compound,  if possible
                       (i.e., unknown aromatic,  unknown hydrocarbon,  unknown
                      acid type,  unknown chlorinated compound).   If probable
                      molecular weights  can be  distinguished,  include them  .

 11.1.2.5.7         The Contractor shall  report  pesticide target compounds
                   listed in Exhibit C (Pesticides) that appear as semivolatile
                   tentatively identified compounds.
                                    D-46/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles -- Section 11
                                                 Data Analysis and Calculations
                                                                    Calculations

11.1.2.5.8        Straight-chain, branched, or cyclic alkanes are   not  to be
                  reported as tentatively  identified compounds on  FORM  I  SV-
 —  -  -       -  TIC.- When the above-alkanes are- tentatively--identified,  the
                  concentration(s) are  to  be  estimated as described in  Section
                  11.2.2 and reported in the  SPG Narrative  as  alkanes.  bv
                  class  (i.e.,  straight-chain, branched, or cyclic;  as  a  "
                  series; as applicable).

11.2  Calculations

11.2.1   Target Compounds

11.2.1.1       Target  compounds identified shall  be quantitated  by the
               internal standard method.   The internal standard  used shall be
               the one assigned to that analyte  for quantitation (see Table
               2).  The EICP area of  primary  characteristic ions of  analytes
               listed  in Tables 3 and 4 are used  for quantitation.

11.2.1.2       It is expected that situations will  arise when the  automated
               quantitation procedures  in  the GC/MS software provide
               inappropriate quantitations.   This normally  occurs  when  there
               is compound co-elution,  baseline noise, or matrix
               interferences.   In these circumstances, the  Contractor must
               perform a manual quantitation.  Manual quantitations  are
               performed by integrating the area  of the quantitation ion  of
               the compound.  This integration shall only include  the area
               attributable to  the specific TCL  compound.   The area integrated
               shall not include baseline  background noise.  The area
               integrated shall not extend past  the point where  the  sides of
               the peak intersect with  the baseline noise.   Manual integration
               is not  to be used solely to meet QC  criteria, nor is  it  to be
               used as a substitute for corrective  action on the
               chromatographic  system.   Any instance of manual integration
               must be documented in  the SDG  Narrative.

11.2.1.3       In all  instances where the  data system report has been edited
               or where manual  integration or quantitation  has been performed,
               the GC/MS operator must  identify  such edits  or manual
               procedures by initialing and dating  the changes made  to  the
               report, and shall include the  integration scan range.  In
               addition, a hardcopy printout  of  the EICP of the  quantitation
               ion displaying the manual integration shall  be included  in the
               raw data.  This  applies  to  all compounds listed in  Exhibit C
                (Semivolatiles) , internal standards, and surrogates.

11.2.1.4       The requirements listed  in  Sections  11.2.1.2 and  11.2.1.3  apply
               to all  standards, samples including  MS/MSD,  and blanks.

11.2.1.5       The relative response  factor (RRF)  from the  continuing
               calibration analysis is  used to calculate the concentration in
               the sample.  For samples analyzed  during the same 12-hour  time
               period  as the initial  calibration  standards,  use  the  RRF values
               from the 50 ng calibration  standard.  Secondary ion
               quantitation is  allowed  ONLY when  there are  sample

                                   D-47/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 11
Data Analysis and Calculations
Calculations

                interferences with the  primary ion.   If secondary ion
                quantitation  is  performed,  document  the reason in the SDG
                Narrative.  The  area  of a secondary  ion cannot be used for the
                area of  a primary ion unless a relative response factor is
                calculated  using the  secondary ion.

11.2.1.6        Calculate the concentration in the sample using the relative
                response factor  (RRF) and the following equations.

11.2.1.6.1         Water

                   EQ.   5

                                         (A ) (I )  (V ) (Df) (GPC)
                   Concentration ug/L = 	;—:—
                                 ^       (Ais) (RRF)  (Vo)  (V±)

                   Where,

                   ~AX    =  Area of  the  characteristic ion for  the compound to
                            be measured.'
                   Als    =  Area of  the  characteristic ion for  the internal
                            standard.
                   IB    =  Amount of  internal standard  injected  in  nanograms
                             (ng) .
                   V0    =  Volume of  water  extracted  in milliliters (mL).
                   Vt    =  Volume of  extract injected in microliters  (uL).
                   Vt    =  Volume of  the concentrated extract  in microliters
                             (uL)  (Vt = 1,000 uL  if sample was not subjected to
                            GPC; Vt  =  500 uL if  sample was subjected to GPC).
                   RRF   =  Relative response factor determined from the  12-
                            hour calibration standard.
                   GPC   =  GPC  factor.
                            GPC  = 1.0  if water sample  was not subjected  to GPC;
                            GPC  = 2.0  if water sample  was subjected  to GPC.
                   Df    =  Dilution factor.  The dilution factor for analysis
                            of water samples for semivolatiles  by this method
                             is defined as  follows:
          , _ uL most cone,  extract used to make dilution  + uL clean solvent
                      uL most cone,  extract used to make  dilution
                          If  no dilution  is performed, Df  =  1.0.
                                    D-48/SVOA                             OLM04.2

-------
11.2.1.6.2
                                         Exhibit D Semivoiatiies -- Section 11
                                                Data Analysis and Calculations
                                                                  Calculations
Soil/Sediment

EQ.  6 	
             Concentration ug/Kg  (Dry weight basis) =
                                   (Ax)
                                                              (Vt) (Df)  (GPC)
                                                      (A.s)  (RRF) (V.) (WJ (D)
Where,

**x'  ^s
vt

Vi

D
                  GPC
                  RRF

                  Df
                         and Ais,  are as given for water,  above.
                           Volume of the concentrated extract in microliters
                            (uL)  (Vt = 500 uL).
                           Volume of the extract  injected in microliters  (uL)
                            100 - % Moisture
                 100

         Weight of sample extracted in grams  (g)
         GPC factor  (GPC = 2.0 to account for GPC cleanup).
         Relative response factor determined from the 12-
         hour calibration standard.
         Dilution factor.  The dilution factor for analysis
         of soil/sediment samples for semivolatiles by this
         method is defined as follows:
              _ _ uL most cone, extract used to make dilution + uL clean solvent
                          uL most cone.  extract used to make dilution
                  If no dilution is performed, Df = 1.0.

                  The factor of 2.0 in the numerator is used to account for
                  the amount of extract that  is not recovered from the
                  mandatory use of GPC cleanup.  Concentrating the extract
                  collected after GPC to O.5  mL maintains the sensitivity of
                  the soil/sediment method.

11.2.2   Non-Target Compound

         An estimated concentration for non-target compounds tentatively
         identified shall be quantitated by the internal standard method.  For
         quantitation, the nearest internal standard free of interferences
         shall be used.  The equations for calculating concentration are the
         same as Equations 5 and 6.  Total area counts  (or peak heights) from
         the total ion chromatograms are to be used for both the compounds to
         be measured and the internal standard.  A relative response factor
         (RRF) of one  (1) is to be assumed.   The resulting concentration shall
         be qualified as "J"  (estimated, due  to lack of a compound specific
         response factor), and "N"  (presumptive evidence of presence),
         indicating the quantitative and qualitative uncertainties associated
         with this non-target component.  An  estimated concentration should be
         calculated for all tentatively identified compounds as well as those
         identified as unknowns.
                                   D-49/SVOA                     .      OLM04.2

-------
Exhibit D Semivolatiles -- Section 11
Data Analysis and Calculations
Calculations

11.2.3   CRQL Calculations

         If the CRQL  is less  than the  CRQL in Exhibit  C  (Seraivolatiles),
         report the CRQL  listed  in Exhibit C (Semivolatiles).
11.2.3.1        Water Samples
                EQ.   7
                    Adjusted _ Contract    (Vx) (Vt)  (Vy) (Df )
                      CRQL       CRQL   X    (V)  (V) (V)
                Where,

                Vc, Df,  V0, and V\ are as given in Equation 6.
                Vx  =  Contract sample volume  (1,000 mL).
                Vy  =  Contract injection volume  (2 uL).
                Vc  =  Contract concentrated extract volume (1,000 uL if GPC is
                      not performed, and 500 uL  if GPC was performed.
 11.2.3.2        Soil/Sediment Samples
                EQ.  8
                    Adjusted  _ Contract    (Wx) (Ve) (Vy) (Df)
                       CRQL       CRQL      (W(j) (vj  (V± )  (D)
                Where,

                Vc,  Df,  Ws,  Vi( and D are as given in Equation 6.
                Wx =  Contract sample weight  (30 g  for low  level and  1  g  for
                      medium  level soil/sediment samples).
                Vy =  Contract injection volume  (2  uL).
                Vc =  Contract concentrated extract volume  (500 uL, GPC is
                      required).
 11.2.4   Surrogate Recoveries
 11.2.4.1       Calculate surrogate standard recovery on all  samples, blanks,
                matrix spikes, and matrix  spike duplicates.   Determine  if
                recovery is within limits  (see Table 7) and report  on
                appropriate form.
                                    D-50/SVOA                             OLM04.2

-------
                                          Exhibit D Semivolatiles -- Section 11
                                                 Data Analysis and Calculations
                                                                    Calculations

11.2.4.2       Calculate the concentrations of the surrogate compounds using
               the same equations as used for the target compounds.  Calculate
               the recovery of each surrogate-using, the following- equation:
               EQ.  10
               . _           Concentration (or amount)  found   , „„
               % Recovery = 	:	 x 100
                            Concentration  (or amount) spiked
11.3  Technical Acceptance Criteria' for Sample Analysis

11.3.1   The samples must be analyzed on a GC/MS system meeting the DFTPP
         initial calibration, continuing calibration, and blank technical
         acceptance criteria.  the sample must undergo cleanup procedures,
         when required, on a GPC meeting the technical acceptance criteria for
         GPC calibration.

11.3.2   The sample must be extracted-and analyzed within the contract holding
         times.

11.3.3   The sample must have an associated method blank meeting the blank
         technical acceptance criteria.

11.3.4   The sample must have an acceptable surrogate recovery.  The sample
         surrogate recovery will be considered unacceptable if the following
         occurs.

         •  Any surrogate has a recovery less than 10.0 percent.

         •  More than  one base neutral surrogate or more than one acid
            surrogate  is outside the  surrogate acceptance windows listed  in
            Table 7.

         NOTE:  These  requirements do not apply to surrogates with advisory
         windows and to dilutions of  samples.

11.3.5   The relative  retention time  of each surrogate must be within ±0.06
         RRT units of  its relative retention time in the continuing
         calibration standard.

11.3.6   The instrumental response  (ElCP.area) for each of the internal
         standards must be within the inclusive range of -50.0 percent and
         +100.0 percent of•the response of the internal standards in the  most
         recent continuing calibration analysis.

11.3.7   The retention time shift for each of the internal standards must be
         within ± 0.50 minutes  (30 seconds) between the sample and the most
         recent continuing calibration standard analysis.

11.3.8   Excluding those ions in the  solvent front, no ion may saturate the
         detector.  No target compound concentration may exceed the upper

  '  •'. •                             D-51/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 11
Data Analysis and Calculations
Corrective Action

         limit of the 12-hour standard calibration range unless a more dilute
         aliquot of the sample extract is also analyzed according to the
         procedures in Section 10.6.5.

11.4  Corrective Action

11.4.1   Corrective Action for Sample Analysis

         The sample technical acceptance criteria  must be met before data are
         reported.  Samples  contaminated from laboratory sources, or sample
         results submitted not meeting the sample technical acceptance
         criteria, will require  re-extraction and/or reanalysis at no
         additional, cost  to  the  Agency.

11.4.2   Corrective Action for Surrogate Recoveries which Fail to Meet Their
         Acceptance Criteria (Section 11.3.4, Table 7).

11.4.2.1        If  the  surrogate  compounds  fail  to meet their  recovery
                acceptance criteria,  check  calculations,  sample preparation
                logs,  surrogate  compound  spiking solutions,  and the  instrument
                operation.  If the calculations  were incorrect, correct  the
                calculations  and verify that  the surrogate compound  recoveries
                meet  their acceptance criteria.   If the sample preparation logs
                indicate  that the incorrect amount of  surrogate compound
                spiking solution was  added, then re-extract/reanalyze the
                sample  after  adding the correct  amount  of surrogate  spiking
                solution.   If the surrogate compound spiking solution was
                improperly prepared,  concentrated, or  degraded, re-prepare
                solutions  and re-extract/reanalyze samples.   If the  surrogate
                recoveries were  outside  the lower surrogate  acceptance  limit
                and the extract  from the  sample  was cleaned  up on a  GPC using
                an automated injection system,  the Contractor shall  verify that
                the proper amount was injected on the  GPC column.   If
                insufficient  sample volume  was injected on the GPC,  the sample
                must be reprepared and reanalyzed at no additional cost to the
                Agency.   If the instrument  malfunctioned,  correct the
                instrument problem and reanalyze the  sample  extract.  Verify
                that the surrogate recoveries meet their acceptance  criteria.
                If the instrument malfunction affected the calibrations,
                recalibrate the instrument  before reanalyzing the sample
                extract.

 11.4.2.2        If the above actions do not correct  the problem,  then  the
                problem may be due to a sample matrix  effect.   To determine  if
                there was matrix effect,  take the following corrective  action
                steps.

 11.4.2.2.1        Re-extract and reanalyze the sample.  EXCEPTION:   If
                   surrogate compound recoveries in  a sample used for a matrix
                   spike and/or matrix spike  duplicate were considered
                   unacceptable, then it should be re-extracted/reanalyzed only
                   if surrogate compound recoveries  met the surrogate
                   acceptance criteria in both the matrix spike and matrix
                   spike duplicate analyses.

                                    D-52/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles -- Section 11
                                                 Data Analysis and Calculations
                                                              Corrective Action

11.4.2.2.2        If the surrogate compound  recoveries meet acceptance
                  criteria  in the re-extracted/reanalyzed  sample, then  the
       	 . problem-was-within -the Contractor's-control.   Therefore,
                  submit only data from the  re-extraction/reanalysis.

11.4.2.2.3        If the surrogate compound  recoveries fail to meet the
                  acceptance criteria in the re-extracted/reanalyzed sample,
                  then  submit data from both analyses.   Distinguish between
                  the initial analysis and the  re-extraction/reanalysis on all
                  deliverables, using the suffixes  in Exhibit B.

11.4.3   Corrective Action  for Internal Standard Compound Responses which Fail
         to Meet their Acceptance Criteria (Sections  11.3.5 and  11.3.6) .

11.4.3.1        If the internal standards fail to meet their acceptance
                criteria, check calibrations,  internal standard compound
                spiking  solutions, and instrument operation.  If  the
                calculations were incorrect,  correct  the  calculations and
                verify that  the internal standard response  met their acceptance
                criteria.  If the internal standard  compound spiking solution
                was improperly prepared, concentrated, or degraded, re-prepare
                solutions and re-extract/reanalyze samples.  If the instrument
                malfunctioned correct the instrument  problem and  reanalyze the
                sample extract.  If the instrument malfunction affected  the
                calibration, recalibrate the  instrument before reanalyzing the
                sample extract.

11.4.3.2        If the above actions do not correct  the problem,  then the
                problem  may  be due to a sample matrix effect.  To determine if
                there was matrix effect, take the following corrective action
                steps.

11.4.3.2.1        Reanalyze the sample extract.  EXCEPTION:   If  internal
                  standard  compound recoveries  in a sample used  for a matrix
                  spike and/or matrix spike  duplicate were outside  the
                  acceptance windows, then it should be  reanalyzed only if
                  internal  standard compound recoveries  met the  internal
                  standard  acceptance criteria  in both the matrix spike and
                  matrix spike duplicate analysis.

11.4.3.2.2        If the internal standard compound recoveries meet acceptance
                  criteria  in the reanalyzed sample extract,  then the problem
                  was within  the Contractor's control.   Therefore,  submit only
                  data  from the reanalysis.

11.4.3.2.3        If the internal standard compound recoveries fail to  meet
                  their acceptance windows in the reanalyzed  sample extract,
                  then  submit data from both analyses.   Distinguish between
                  the initial analysis and the  reanalysis  on  all deliverables
                  using the suffixes in Exhibit B.

11.4.4   Corrective Action  for Surrogate Compound Relative Retention
         Times/Internal Standard Compound Retention  Times  Outside Acceptance
         Criteria (Sections 11.3.5 and 11.3.6).

                                   D-53/SVOA                           OLM04.2

-------
Exhibit D Semivolatiles -- Section 11
Data Analysis and Calculations
Corrective Action

11.4.4.1        If the  surrogate  compounds relative retention  times or  internal
                standard compounds  retention  times are not within  their
                acceptance criteria,  check the  instrument for  malfunctions.   If
                the  instrument  malfunctioned, correct the instrument problem
                and  reanalyze the sample  extract.  If the instrument
                malfunction affected  the  calibration, recalibrate  the
                instrument before reanalyzing the sample extract.

11.4.4.2        If the  above actions  do not correct the problem, then the
                problem may be  due  to a sample  matrix effect.   To  determine if
                there was  matrix  effect,  take the following  corrective  action
                steps.

11.4.4.2.1        Reanalyze the  sample extract.  EXCEPTION:  If the  surrogate
                  compounds relative retention times or  internal  standard
                  compounds retention times  in a sample  used for  a matrix
                  spike  and/or matrix spike  duplicate were  outside the
                  acceptance  criteria, then  it should be reanalyzed  only if
                  the  surrogate  compounds  relative  retention times and
                   internal standard compounds  retention  times were within the
                  acceptance  criteria in both  the matrix spike and matrix
                   spike  duplicate  analysis.

11.4.4.2.2         if the surrogate compounds relative  retention times  and
                   internal standard compounds  retention  times are within the
                   acceptance criteria in the reanalyzed  sample extract,  then
                   the problem was  within the Contractor's control.  Therefore,
                   submit only data from  the  reanalysis with the surrogate
                   compounds relative retention times  and the internal  standard
                   compound retention times within  the  acceptance limits.

11.4.4.2.3         If the surrogate compounds relative  retention times  or
                   internal standard compounds  retention  times are outside the
                   acceptance criteria in the reanalyzed  sample extract,  then
                   submit data from both analyses.   Distinguish between the
                   initial analysis and the reanalysis  on all deliverables,
                   using the suffixes in Exhibit B.

11.4.5   Corrective action for failure to  meet instrument performance  checks
          and initial .and continuing calibration must be  completed before the
          analysis of samples.
                                    D-54/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles -- Section 12
                                                                Quality Control
                                                                  Method Blanks

12.0  QUALITY CONTROL

12.1  Method Blanks -    ._.-__  . .  .    __    _ .

12.1.1   Summary of Method Blanks

         A method blank  is a volume of a clean reference matrix  (reagent water
         for water samples, or purified sodium sulfate for soil/sediment
         samples) that is carried through the entire analytical procedure.
         The volume or weight of the reference matrix must be approximately
        • equal to the volume of weight of samples associated with the blank.
         The purpose of  a method blank is to determine the levels of
         contamination associated with the processing and analysis of samples.

12.1.2   Frequency of Method Blanks

         A method blank  must be extracted each time samples are extracted.
         The number of samples extracted with each method blank shall not
         exceed 20 field samples  (excluding matrix spikes/matrix spike
         duplicates and  PE samples).  In addition, a method blank shall:

         •     Be extracted by  the  same procedure used to extract samples.

         •     Be analyzed on each  GC/MS  system  used to analyze  associated
               samples.

12.1.3   Procedure for Method Blank Preparation

         For semivolatile analyses, a method blank for water samples consists
         of 1 L volume of reagent water spiked with 0.5 mL of the surrogate
         spiking solution.  For medium and low level soil/sediment samples, a
         method blank consists of 1 g and 30 g of sodium sulfate spiked with
         0.5 mL of the surrogate spiking solution respectively.  Extract,
        . concentrate, cleanup, and analyze the blank according to procedures
         for water and soil samples.

12.1.4   Technical Acceptance Criteria for Method Blank Analysis

12.1.4.1       All blanks must be extracted and  analyzed at the  frequency
               described in Section 12.1.2 on a  GC/MS system meeting the
               DFTPP, initial calibration, and continuing calibration
               technical acceptance criteria.

12.1.4.2       The blank must meet  the sample acceptance criteria listed in
               Sections  11.3.4  through 11.3.7.

12.1.4.3       For all other target compounds the method blank must contain
               less  than the CRQL of any  single  target compound  (see Exhibit
               C).  A method blank  for semivolatile analysis must contain less
               than  five times  (5x) the contract required quantitation limit
                (CRQL) of the phthalate esters listed in Exhibit  C.

12.1.5   Corrective Action for Method Blanks
                                   D-55/SVOA                             OLM04.2

-------
Exhibit D Semivolatiles -- Section 12
Quality Control
Matrix Spike/Matrix Spike Duplicate  (MS/MSD)

12.1.5.1        If a method blank  does not meet the technical  acceptance
                criteria  for  method blank analysis, the  Contractor  shall
                consider  the  analytical  system to be out of  control.

12.1.5.2        If contamination is the  problem, then  the source  of the
                contamination must be investigated and appropriate  corrective
                measures  must be taken and documented  before further sample
                analysis  proceeds. It is the Contractor's responsibility to
                ensure  that method interferences caused  by contaminants in
                solvent,  reagents, glassware, and sample storage  and processing
                hardware  that lead to discrete artifacts and/or elevated
                baselines in  the GC/MS be eliminated.  Samples associated with
                the contaminated blank must  be re-extracted  and reanalyzed at
                no additional cost to the Agency.

12.1.5.3        If surrogate  recoveries  in the method  blank  do not  meet the
                acceptance criteria listed in Section  11.3.4 and  Table 7,  first
                reanalyze the method  blank.  If the surrogate recoveries
                (except those with advisory  limits) do not-meet the acceptance
                criteria after reanalysis, the method  blank  and all samples
                associated with that  method  blank must be re-extracted and
                reanalyzed at no additional  cost  to the  Agency.

12.1.5.4        If the  method blank does not meet  internal standard response
                requirements  listed in Section  11.3.6, check calculations, the
                internal standard spiking solutions, and the instrument
                operation.  If the calculations were incorrect, correct  the
                calculations  and verify  that the  internal standard responses
                meet  their acceptance criteria.   If the  internal  standard
                compound spiking solution was  improperly prepared,
                concentrated, or degraded,  re-prepare  solutions and re-
                extract/reanalyze samples.   If  the  instrument malfunctioned,
                correct the instrument problem  and reanalyze the  method  blank.
                If  the  instrument malfunction  affected the calibration,
                recalibrate the instrument before reanalyzing the blank.   The
                Contractor shall resolve and document  the resolution of  the
                problem before proceeding with  sample  analysis.

12.1.5.5        If  the  method blank does not meet the  retention time
                requirements for  internal standards (Section 11.3.7) or  the
                surrogates (Section 11.3.5), check the instrument for
                malfunction and recalibrate.  Reanalyze  the method blank.
                Sample  analyses cannot proceed until  the method blank meets
                these requirements.

12.2   Matrix Spike/Matrix Spike Duplicate (MS/MSD)

12.2.1   Summary of  MS/MSD

          In order  to evaluate the effects of  the sample matrix on the methods
          used for  semivolatile analyses,  the Agency has prescribed a mixture
          of semivolatile target  compounds to  be  spiked  into two  aliquots of  a
          sample and analyzed in accordance with the  appropriate  method.
                                    D-5S/SVOA                            OLM04.2

-------
                                          Exhibit D Serr.ivolatiles -- Section 12
                                                                 Quality Control
                                   Matrix Spike/Matrix Spike Duplicate (MS/MSD)

12.2.2   Frequency of MS/MSD Analyses

12.2.2.1       A matrix  spike and matrix  spike duplicate must be extracted  and
               analyzed  for every 20 field  samples of  a similar  matrix in an
               SDG whenever samples are extracted by the same procedure  (  low
               or medium level  soil preparation) . ' MS/MSD samples shall be
               analyzed  unless  otherwise  specified on  the  Traffic Report  (TR).
               If no MS/MSD samples are specified on the TR, the Contractor
               shall contact SMO to confirm that MS/MSD analyses are  not
               required.

12..2.2.2       As part of the Agency's QA/QC program,  water  rinsate samples
               and/or field/trip blanks  (field QC) may accompany soil/sediment
               samples and/or water samples that are delivered to the
               laboratory for analysis.   The Contractor shall not perform
               MS/MSD analysis  on any of  the field QC  samples.

12.2.2.3       If the EPA Region designates a sample to be used  as  an MS/MSD,
               then that sample must be used.  If there is  insufficient sample
               remaining to perform an MS/MSD, then the Contractor  shall
               choose another sample on which to perform an  MS/MSD  analysis.
               At the time the  selection  is made, the  Contractor shall notify
               the Region (through SMO) that insufficient  sample was  received
               and identify the EPA sample  selected for the  MS/MSD  analysis.
               The rationale for the choice of another sample other than the
               one designated by the Agency shall be documented  in  the SDG
               Narrative.

12.2.2.4       If there  is insufficient sample remaining in  any  of  the samples
               in an SDG to perform an MS/MSD, then the Contractor  shall
               immediately contact SMO to inform  them  of the problem.  SMO
               will contact the Region for  instructions.   The Region  will
               either approve that no MS/MSD is required,  or require  that a
               reduced sample aliquot be  used for the  MS/MSD analysis '.  SMO
               will notify the  Contractor of the  Region's  decision.   The
               Contractor shall document  the decision  in the SDG Narrative.

12.2.2.5       If it appears that  the Region has  requested MS/MSD analysis  at
               a greater frequency than required  by the contract, then the
               Contractor shall contact SMO. SMO will contact the  Region to
               determine which  samples should have an  MS/MSD performed on
               them.  SMO will  notify the Contractor of the  Region's  decision.
               The Contractor shall document the  decision  in the SDG
               Narrative.  If this procedure is not  followed,  the  Contractor
               will not  be paid for MS/MSD  analysis performed at a  greater
               frequency than required by the contract.'

12.2.2.6       When a Contractor receives only a  performance evaluation  (PE)
               sample(s), no MS/MSD shall be performed within that  SDG.

12.2.2.7       When a Contractor receives a performance evaluation  (PE) sample
               as part of a larger SDG, a sample  other than  the  PE  sample must
               be chosen for the MS/MSD when the  Region did not  designate
               samples to be used  for this  purpose.   If the  PE sample is an

                                   D:57/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 12
Quality Control
Matrix Spike/Matrix Spike Duplicate  (MS/MSD)
                ampulated standard,  the  ampulated PE  sample  is not  considered
                to be  another matrix type.

12.2.3   Procedure  for  Preparing MS/MSD

12.2.3.1        Water  Samples

                For  water samples,  prepare  two additional  1  L aliquots  of  the
                sample chosen for  spiking  in two continuous  extractors.  Add
                0.5  mL of surrogate spiking solution  and 0.5 mL  of  the  matrix
                spiking solution to each aliquot.  Extract,  concentrate,
                cleanup, and analyze the MS/MSD according to the procedures for
                water  samples.

12.2.3.2        Soil/Sediment Samples -  Low Level

                For  low level soil/sediment samples,  prepare two additional 30
                g aliquots (record weight to nearest  0.1 g)  of  the  sample
                chosen for spiking in the two 400 mL  beakers. Add  60 g of
                anhydrous powdered sodium sulfate to  each aliquot.   Mix well.
                Add  0.5 mL of the  surrogate standard'spiking solution and  0.5
                mL of  the matrix spiking solution to  each aliquot,  then
                immediately add 100 mL of 1:1 methylene chloride-acetone.
                Extract, concentrate, cleanup, and analyze the  MS/MSD according
                to the procedures  for low level soil  samples.

12.2.3.3        Soil/Sediment Samples -  Medium Level

                For  medium level soil/sediment samples, prepare two additional
                1 g  aliquots  (record weight to nearest 0.1 g) of the sample
                chosen for spiking in two 20 mL vials.  Add 2.0 g of anhydrous
                powdered sodium sulfate to each aliquot.  Mix well.  Add 0.5 mL
                of surrogate spiking solution, 0.5 mL of matrix spiking
                solution, and 9.0  mL of methylene chloride.   Extract,
                concentrate, cleanup, and analyze the MS/MSD according to the
                procedures for medium level samples.

12.2.4   Dilution of  MS/MSD

          Before any MS/MSD analysis, analyze the original sample,  then analyze
          the MS/MSD at the same concentration as the most concentrated extract
          for which the original sample results will  be reported.  For example,
          if the original sample is to be reported at a 1:1 dilution and a 1:10
          dilution,  then analyze and report the MS/MSD at a 1:1 dilution only.
          However,  if the original sample is to be reported at a 1:10 dilution
          and a 1:100 dilution, then the MS/MSD must be analyzed and reported
          at a 1:10  dilution only.  Do  not further dilute the MS/MSD samples  to
          get either spiked or non-spiked analytes within calibration range.
          Dilution of the sample must be performed in accordance to the
          conditions in Section 10.6.5.
                                    D-58/SVOA                            OLM04.2

-------
                                          Exhibit D Semivolatiles -- Section 12
                                                                 Quality  Control
                                   Matrix Spike/Matrix Spike Duplicate (MS/MSD)
12.2.5   Calculations for MS/MSD
12.2.5.1       Calculate  the  recovery of  each matrix  spike compound  in  the
               matrix  spike and matrix  spike duplicate and report on the
               appropriate forms.  Calculate the  concentrations of the  matrix
               spike compounds using the  same equations as used for  target
               compounds  (Equations 6 and 7).  Calculate  the recovery of  each
               matrix  spike compound as follows:
               EQ.   11
                                             SSR ~ SR
                     Matrix Spike Recovery = 	 x 100
                                                SA
               Where,

               SSR    =  Spike Sample Result.
               SR    =  Sample Result.
               SA    =  Spike Added.

12.2.5.2       Calculate  the  relative percent  difference  (RPD) of the
               recoveries of  each  compound  in  the matrix  spike and matrix
               spike  duplicate as  follows:
                           RPD =   IMSR-MSDRI  x 100
               EQ.  12

                           RPD =
                                 -j (MSR + MSDR)



               Where,

               RPD    =   Relative  Percent  Difference.
               MSR    =   Matrix Spike Recovery.
               MSDR   =   Matrix Spike Duplicate Recovery.

               The vertical bars  in the formula above indicate the absolute
               value  of  the difference, hence  RPD is always expressed as a
               positive  value.

12.2.6   Technical Acceptance  Criteria for MS/MSD

12.2.6.1       All MS/MSD must be prepared and analyzed at a frequency
               described in Section 12.2.2.  All MS/MSD must be analyzed on a
               GC/MS  system meeting DFTPP, initial and continuing calibration
               technical acceptance criteria,  and the method blank technical
               acceptance criteria.  The  MS/MSD must undergo cleanup
               procedures, when required  on a  GPC meeting the technical
               acceptance criteria for GPC calibration.

12.2.6.2       The MS/MSD must have an associated method blank meeting the
               blank  technical acceptance criteria.

                                   D-59/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 12
Quality Control
Matrix Spike/Matrix Spike Duplicate  (MS/MSD)

12.2.6.3       The MS/MSD must be extracted and  analyzed within  the contract
               holding  time.

12.2.6.4       The retention  time shift  for each of  the internal standards
               must be  within ±0.50  minutes (30  seconds) between the  MS/MSD
               sample and the most recent  continuing calibration standard.

12.2.6.5       The relative retention time for the surrogate  must be  within  ±
               0.06 RRT units of  its standard retention time  in  the continuing
               calibration  standard.

12.2.6.6       The limits for matrix spike compound  recovery  and RPD  are  given
               in Table 6.  As these limits are  only advisory, no further
               action by the  laboratory  is required; however,  frequent
               .failures to  meet  the  limits for recovery or RPD warrant
               investigation  by  the  laboratory,  and  may result in questions
               from the Agency.

12.2.7    Corrective Action  for MS/MSD

          Any MS/MSD which fails  to meet  the technical acceptance criteria for
          MS/MSD must  be reanalyzed at no additional  cost to the  Agency.
                                    D-60/SVOA                            OLM04.2

-------
                                      Exhibit D Semivolatiles -- Sections 13-16
           Method Performance/Pollution Prevention/Waste Management/References

13.0  METHOD PERFORMANCE

      Not Applicable.

14.0  POLLUTION PREVENTION

14.1  Pollution prevention encompasses any technique that reduces or
      eliminates the quantity or toxicity of waste at the point of generation.
      Numerous opportunities for pollution prevention exist  in laboratory
      operation.  The EPA has' established a preferred hierarchy of
      environmental management techniques that places pollution prevention as
      the management option of first choice.  Whenever feasible, laboratory
      personnel should use pollution prevention techniques to address their
      waste generation.  When wastes cannot be feasibly reduced at the source,
      the Agency recommends recycling as the next best option.

14.2  For information about pollution prevention that may be applicable  to
      laboratories and research institutions, consult Less is Better:
      Laboratory Chemical Management for Waste Reduction, available from the
      American Chemical Society's Department of Government Relations and
      Science Policy, 1155 16th Street, N.W., Washington D.C.  .20036,  (202)
      872-4477.

15.0  WASTE MANAGEMENT

      The Environmental Protection Agency requires that laboratory waste
      management practices be conducted consistently with all applicable rules
      and regulations.  The Agency urges laboratories to protect the air,
      water, and land by minimizing and controlling all releases from hoods
      and bench operations, complying with"the letter and spirit of any  sewer
      discharge permits and regulations, and by complying with all solid and
      hazardous waste regulations, particularly the hazardous waste
      identification rules and land disposal restrictions.   For further
      information on waste management consult The Waste Management Manual for
      Laboratory Personnel, available from the American Chemical Society at
      the address listed in Section 14.

16.0  REFERENCES

      Not Applicable.
                                    D-61/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Section 17
Tables/Diagrams/Flowcharts

17 . 0  TABLES/DIAGRAMS/FLOWCHARTS


                                    Table  1


                   DFTPP Key Ions and Ion. Abundance Criteria

                Mass       Ion Abundance  Criteria



                  51       30.0-80.0 percent  of  mass  198

                  68       Less  than  2.0  percent of  mass  69

                  69       Present

                  70       Less  than  2.0  percent of  mass  69

                 127       25.0  -  75.0 percent  of  mass  198

                 197       Less  than  1.0  percent of  mass.198

                 198       Base  peak  100  percent relative
                           abundance  (see Note)

                 199       5.0  - 9.0  percent  of mass 198

                 275       10.0  -  30.0 percent  of  mass  198

                 365       Greater than  0.75  percent of mass  198

                 441       Present but less than mass 443

                 442       40.0  -  110.0  percent of mass 198

                 443       15.0  -  24.0 percent  of  mass  442

       NOTE:   All ion abundances MUST be  normalized  to  m/z 198,  the nominal
       base peak, even though the-ion abundance of m/z  442 may be up to 110
       percent that of m/z 198.
                                    D-62/SVOA                            OLM04.2

-------
                                               Exhibit D Semivolatiles -- Section 17
                                                          Tables/Diagrams/Flowcharts
                                                        Table 2

                                   Semivolatile Internal Standards With Corresponding
                               Target Compounds and Surrogates Assigned for Quantitation
1.4-Dichlorobenzene-d,,
Benzaldehyde
Phenol
bis-(2-Chloroethyl
   ether
2-Chlorophenol
2-Methylphenol
2,2'-oxybis-(1-
Chloropropane)
Acetophenone
,4-Methylphenol
N-Nitroso-di-n-        |
   propylamine
Hexachloroethane
2-Fluorophenol  (surr)
Phenol-d5  (surr)
2-Chlorophenol-d, (surr)
1,2-Dichlorobenzene-d4
   (surr)
Naphthalene-dn
Nitrobenzene
Isophorone
2-Nitrophenol
2,4-Dimethylphenol
bis-(2-Chloro
ethoxy)methane
2,4-Dichlorophenol|
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
Caprolactam
4-Chloro-3-
methylphenol
2-Methylnaphthalene
Nitrobenzene-d5
(surr)
Acenaphthene-din   |
Hexachlorocyclo-
   pentadiene
2,4,6-Trichloro
   phenol
2,4,5-Trichloro
   phenol
1,1'-Biphenyl
2-Chloro-         |
   naphthalene
2-Nitroaniline
Dimethylphthalate
2,6-Dinitrotoluene
Acenaphthylene
3-Nitroaniline
Acenaphthene
2,4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2,4-Dinitrotoluene
Diethylphthalate
Fluorene.
4 -Chlorophenyl-
phenylether
4-Nitroaniline
2 -Fluorobiphenyl
(surr)
Phenanthrene-d, „
4,6-Dinitro-2-
methylphenol
4-Bromophenyl-
phenylether
N-Nitroso
diphenylamine
Hexachlorobenzene
Atrazine
Pentachlorophenol
Phenanthrene
Anthracene
Carbazole
Di-n-butylphthalate
Fluoranthene
2,4,6-Tri-
Bromophenol
(surr)
Chrvsene-d,,
Pyrene
Butylbenzyl
phthalate
3,3'-Dichloro
benzidine
Benzo(a)anthracene
Chrysene
bis-(2-Ethylhexyl)
   phthalate
Terphenyl-d14  (surr)
Pervlene-d,,
Di-n-octyl
phthalate
Benzo(b)
   fluoranthene
Benzo(k)
   fluoranthene
Benzo(a)pyrene
Indenod, 2, 3-cd)
   pyrene
Dibenzo(a,h)-
anthracene
Benzo(g,h,i)
perylene
surr = surrogate  compound
                                        D-63/SVOA
                                                    OLM04.2

-------
Exhibit D Semivolatiles -- Section 17
Tables/Diagrams/Flowcharts
                                     Table  3

                   Characteristic  Ions  for  Internal  Standards
                           for Semivolatile Compounds

                                           Primary
                                         Quantitation
          INTERNAL STANDARDS                 Ion        Secondary  Ions

           l,4-Dichlorobenzene-d4            152        115

           Naphthalene-d8                    136        68

           Acenaphthene-dlO                  164        162, 160

           Phenanthrene-dlO                  188        94,  80

           Chrysene-dl2                      240        120, 236

           Perylene-dl2                      264  .      260, 265
                                    D-64/SVOA                            OLM04.2

-------
                    Exhibit D Semivolatiles -- Section 17
                                Tables/Diagrams/Flowcharts
               Table 4
Characteristic Ions for Semivolatile
   Target  Compounds  and Surrogates
Parameter
Benz aldehyde
Phenol
bis- (2-Chloroethyl) ether
2 -Chlorophenol
2 -Methylphenol
2,2' -oxybis (1-Chloropropane)
Acetophenone
4 -Methylphenol
N-Nitroso-di-n-propylamine
Hexachloroethane
Nitrobenzene
Isophorone
2 -Nitrophenol
2 , 4-Dimethylphenol
bis- (2 -Chloroethoxy) methane
2 , 4-Dichlorophenol
Naphthalene
4 -Chloroaniline
Hexachlorobutadiene
Caprolactam
4 - Chloro- 3 -methylphenol
2 -Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6-Trichlorbphenol
2,4, 5-Trichlorophenol
1,1' -Biphenyl
2 -Chloronaphthalene
2 -Nitroaniline
Dimethylphthalate
2 , 6 -Dinitrotoluene
Acenaphthylene
3 -Nitroaniline
Acenaphthene
2 , 4 -Dinitrophenol
4 -Nitrophenol
Dibenzofuran
2 , 4 -Dinitrotoluene
Diethylphthalate
Fluorene
4-Chlorophenyl-phenylether
4 -Nitroaniline
4 , 6-Dinitro-2-methylphenol
4-Bromophenyl-phenylether
N-Nitrosodiphenylamine
Hexachlorobenzene
Atrazine
Pentachlorophenol
Primary
Quant it at ion
Ion
77
94
93
128
108
45
105
108
70
117
77
82
139
107
93
162
128
127
225
113
107
142
237
196
196
154
162
65
163
165
152
138
153
184
109
168
165
149
166
204"
138
198
248
169
284
200
266
Secondary Ion(s)
105, 106
65, 66
63, 95
64, 130
107
77, 79
77, 51
107
42, 101, 130
201, 199
123, 65
95, 138
65, 109
121, 122
95, 123
164, 98
129, 127
129
223, 227
55, 56
144, 142
141
235, 272
198, 200
198, 200 .
153, 76
164, 127
92, 138
194,- 164
89, 121
151, 153
108, 92
152, 154
63, 154
139, 65
139
63, 182
177, 150
165, 167
206, 141
92, 108
182, 77
250, 141
168, 167
142, 249
173, 215
264, 268
              D-65/SVOA
OLM04.2

-------
Exhibit D Semivolatiles -- Section 17
Tables/Diagraihs/Flowcharts
                                Table 4 (cont'd)
                      Characteristic Ions  for Semivolatile
                        Target Compounds and  Surrogates
                                    Primary
                                    Quantisation

      Parameter                     Ion          __SecgndaryLjIgiils_L
      Phenanthrene                  178             179,  1?6
      Anthracene                    178             179,  176
      Carbazole                     167             166,  139
      Di-n-butylphthalate          149             150,  104
      Fluoranthene                  202             101,  100
      Pyrene                        202             101,  100
      Butylbenzylphthalate         149             91,  206
      3,3'-Dichlorobenzidine       252             254,  126
      Benzo(a)anthracene           228             229,  226
      Chrysene                     228             226,  229
      bis-(2-Ethylhexyl)phthalate  149             167,  279
      Di-n-Octyl  phthalate         149             none
      Benzo(b)fluoranthene         252             253,  125
      Benzo(k)fluoranthene         252             253,  125
      Benzo(a)pyrene               252             253,  125
      Indeno(l,2,3-cd)pyrene       276             138,  227
      Dibenzo(a,h)anthracene       278             139,  279
      Benzo(g,h,i)perylene         276             138,  277
      SURROGATES
      Phenol-d5                      99             42,  71
      2-Fluorophenol                112             64
      2,4,6-Tribromophenol          330             332,  141
      Nitrobenzene-d5               82             128,  54
      2-Fluorobiphenyl              172             171
      Terphenyl-dl4                 244             122,  212
      2-Chlorophenol-d4             132             68,  134
      1. 2-Dinhl orohgri7:eng-d4	152	115.  150
                                    D-66/SVOA                             OLM04.2

-------
                                 Exhibit D Semivolatiles -- Section 17
                                            Tables/Diagrams/Flowcharts

                           Table 5
Relative Response Factor Criteria for Initial and Continuing
 Calibration  of  Semivolatile  Target-Compounds  and Surrogates
Semivolatile
Compounds
Benzaldehyde
Phenol
bis- (2-Chloroethyl) ether
2 -Chlorophenol
2 -Methylphenol
2,2' -oxybis- (1-Chloropropane)
Acetophenone
4 -Methylphenol
N-Nitroso-di-n-propylamine
Hexachloroethane
Nitrobenzene
Isophorone
2 -Nitrophenol
2 , 4 -Dimethylphenol
bis- ( 2 -Chloroethoxy) methane
2 , 4 -Dichlorophenol
Naphthalene
4 -Chloroaniline
Hexachlorobutadiene
Caprolactam
4 - Chloro - 3 - Me thylphenol
2 -Methylnaphthalene
Hexachlorocyclopentadiene
2,4,6- Tri Chlorophenol
2,4, 5 -Tri Chlorophenol
1,1' -Biphenyl
2 -Chloronaphthalene
2-Nitroaniline
Dimethylphthalate
2 , 6 -Dinitrotoluene
Acenaphthylene
3 -Nitroaniline
Acenaphthene
2 , 4 -Dinitrophenol
4 -Nitrophenol
Dibenzofuran
2 , 4 -Dinitrotoluene
Diethylphthalate
4 -Chlorophenyl-phenylether
Fluorene
4 -Nitroaniline
4, 6 -Dinitro- 2 -Methylphenol
4-Bromophenyl-phenylether
N-Nitrosodiphenylamine
Hexachlorobenzene
Atrazine
Pentachlorophenol
Phenanthrene
Minimum
RRF
0.010
0.800
0.700
0.800
0.700
0.010
0.010
0.600
0.500
0.300
0.200
0.400
0.100
0.200
0.300
0.200
0.700
0.010
0.010
0.010
0.200
0.400
0.010
0.200
0.200
0.010
0.800
0.010
0.010
0.200
0.900
0.010
0.900
0.010
0.010
0.800
0.200
0.010
0.400
0.900
0.010
0.010
0.100
0.010
0.100
0.010
0.050
0.700
Maximum
%RSD
none
20.5
20.5
20.5
20.5
none
none
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
none
none
none
20.5
20.5
none
20.5
20.5
none
20.5
none
none
20.5
20.5
none
20.5
none
none
20.5
20.5
none
20.5
20.5
none
none
20.5
none
20.5
none
20.5
20.5
Maximum
%Diff
none
±25.0
±25.0
±25.0
±25.0
none
none
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
none
none
none
±25.0
±25.0
none
±25.0
±25.0
none
±25.0
none
none
±25.0
±25.. 0
none
±25.0
none
none
±25.0
±25.0
none
±25.0
±25.0
none
none
±25.0
none
±25.0
none
±25.0
±25.0
                          D-67/SVOA
OLM04.2

-------
Exhibit D Semivolatiles -- Section 17
Tables/Diagrams/Flowcharts

                               Table 5 (cont'd)
         Relative  Response Factor Criteria  for  Initial  and Continuing
         Calibration of Semivolatile Target Compounds and Surrogates
Semivolatile
Compounds
Anthracene
Carbazole
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Benzo (a) anthracene
Chrysene
bis- (2-Ethylhexyl)phthalate
Di -n-Octylphthalate
Benzo (b) f luoranthene
Benzo (k) f luoranthene
Benzo (a) pyrene
Indeno ( 1 , 2 , 3 - cd ) pyrene
Dibenzo (a, h) anthracene
Benzo (g,h, i)perylene
SURROGATES
Nitrobenzene -d5
2 - Fluorobiphenyl
Terphenyl - d!4
Phenol -d5
2 - Fluorophenol
2,4, 6 -Tribromophenol
2 - Chlorophenol - d4
1 . 2-Dichlorobenzene-d4
Minimum
RRF
0.700
0.010
0.010
0.600
0.600
0.010
0.010
0.800
0.700
0.010
0.010
0.700
0.700
0.700
0.500
0.400
0.500

0.200
0.700
0.500
0.800
0.600
0.010
0.800
0.400
Maximum
%RSD
20.5
none
none
20.5
20.5
none
none
20.5
20.5
none
none
20.5
20.5
20.5
20.5
20.5
20.5

20.5
20.5
20.5
20.5
20.5
none
20.5
20.5
Maximum
%Diff
±25.0
none
none
±25.0
±25.0
none
none
±25.0
±25.0
none
none
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0

±25.0
±25.0
±25.0
±25.0
±25.0
none
±25.0
+25.0
                                    D-68/SVOA
OLM04.2

-------
                                     Exhibit  D Semivolatiles  --  Section  17
                                                Tables/Diagrams/Flowcharts
                               Table 6

                      Matrix Spike Recovery and
                 Relative Percent Difference Limits
Compound
                     %Recovery  RPD
%Recovery  RPD       Soil/       Soil/
Water      Water     Sediment    Sediment
Phenol                        12-110      42       26-90       35

2-Chlorophenol                27-123      40       25-102      50

N-Nitroso-di-n-propylamine    41-116      38       41-126      38

4-Chloro-3-Methylphenol       23-97       42       26-103      33

Acenaphthene                  46-118      31       31-137   .   19

4-Nitrophenol                 10-80       50       11-114      50

2,4-Dinitrotoluene            24-96       38       28-89       47

Pentachlorophenol         .    9-103       50       17-109      47

Pyrene                        26-127      31       35-142      36
                              D-69/SVOA
                                      OLM04.2

-------
Exhibit D Semivolatiles -- Section 17
Tables/Diagrams/Flowcharts
                                     Table 7

                            .Surrogate Recovery Limits
COMPOUND
Nitrobenzene-d5 (Base/Neutral)
2-Fluorobiphenyl (Base/Neutral)
Terphenyl-dl4 (Base/Neutral)
Phenol -d5 (Acid)
2-Fluorophenol (Acid)
2,4,6 -Tribromophenol (Acid)
2-Chlorophenol-d4 (Acid)
1, 2-Dichlorobenzene-d4
(Base/Neutral)
%Recovery
Water
35-114
43-116
33-141
10-110
21-110
10-123
33-110 (advisory)
16-110 (advisory)
%Recovery
Soil/Sediment
23-120
30-115
18-137
24-113
25-121
19-122
20-130 (advisory)
20-130 (advisory)
                                    D-70/SVOA
OLM04.2

-------
                                          Exhibit D Semivolatiles -- Appendix A
                                     Screening of Semivolatile Organic Extracts

APPENDIX A - SCREENING OF SEMIVOLATILE ORGANIC EXTRACTS

1.0   SCOPE AND APPLICATION

1.1   The analytical method described in this section is designed to
      characterize soil/sediment samples from hazardous waste sites for the
      concentration level of organics.

1.2   The method involves the preparation of soil/sediment samples, which may
      contain organic compounds at a level greater than 10,000 ug/kg, and
      screening with a gas chromatograph equipped with a flame ionization
      detector  (GC/FID).

1.3   For soil/sediment samples, the results of the screen are used to
      determine which of the two sample preparation procedures (low or medium)
      is required, and to determine appropriate dilution factor for GC/MS
      analysis.  The results of the screen may also be used to assist the
      analyst in performing Gel Permeation Chromatography  (GPC) cleanup
      procedures on extracts of either water or soil/sediment samples.

1.4   The procedure is designed to allow a quantitation limit for screening
      purposes as low as 10,000 ug/kg for extractable organics.  For analysis
      purposes, the quantitation limit is 10,000 ug/kg for extractable
      organics.  Some samples may contain high concentrations of compounds
      that interfere with the analysis of other components at lower levels;
      the quantitation limits in those cases may be significantly higher.

1.5   These extraction and preparation procedures were developed for rapid and
      safe handling of high concentration hazardous waste samples.  The design
      of the methods thus does not stress efficient recoveries or low limits
      of quantitation of all components.  Rather, the procedures were designed
      to screen, at moderate recovery and sufficient sensitivity, a broad
      spectrum of organic chemicals.  The results of the analyses thus may
      reflect only a minimum of the amount actually present in some samples.

2.0   SUMMARY OF METHOD

2.1   It is mandatory that all soil/sediment samples be characterized as to
      concentration level so that the appropriate analytical protocol is
      chosen to ensure proper quantitation limits for the sample.  Note that
      the terms "low level" and "medium level" are descriptions of the
      concentration ranges that are encompassed by the "low" and "medium"
      level procedures.

2.2   The laboratory is at liberty to select the method of characterization.
      The following two screening methods may be used for soil/sediment sample
      characterization.

      •  Screen an aliquot from the  "low level" 30 g extract or an aliquot
         from the  "medium level" 1 g  extract.

      •  Screen using GC/FID as the screening instrument.
                                    D-71/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Appendix A
Screening of Semivolatile Organic Extracts

2.3   The concentration ranges covered by these two procedures may be
      considered to be approximately 330 ug/kg - 10,000 ug/kg for the low
      level analysis and  >10,000 ug/kg for medium level analysis for
      semivolatile extractables.

2.4   Sample Preparation

2.4.1    Low Level Soil/Sediment

         A 30  g portion of  soil/sediment is mixed with anhydrous powdered
         sodium sulfate and extracted with 1:1 methylene  chloride/acetone
         using an ultrasonic probe.  5.0 mL of the 300 mL (approximate) total
         extract is concentrated to 1.0 mL and screened.   If the original
         sample concentration  is >10,000 ug/Kg the 30 g extract is discarded
         and the medium level  preparation procedure  followed.

2.4.2    Medium Level  Soil/Sediment

         Approximately 1  g  portions of  soil/sediment are  transferred  to vials,
         mixed with anhydrous  powdered  sodium sulfate, and extracted  with
         methylene chloride.  5.0  mL of  the  10.0 mL  extract is concentrated to
         1.0 mL and screened.   If  the -sample  concentration is <10,000 ug/Kg,
         the  1 g extract  is discarded and  the low  level method is  followed.

2.5    GC/FID Screening

       The  concentrated extracts of soil/sediment or  water samples  are screened
       on a gas chromatograph/flame ionization detector using a fused  silica
       capillary column (FSCC)  for  semivolatile priority pollutants and related
       organics.  The  results of these screens will determine whether
       sufficient quantities of pollutants  are present to  warrant analysis  by
       the  medium level protocol.

3.0    INTERFERENCES

       Method  interferences  may be  caused by  contaminants  in solvents,
       reagents, glassware,  and other  sample  processing hardware  that  lead  to
       discrete artifacts  and/or elevated baselines in the chromatograms.   All
       of these materials  routinely must be demonstrated  to be  free from
       interferences under the  conditions of  the  analysis  by running  laboratory
       method  blanks.   Matrix interferences may be  caused by contaminants  that
       are  co-extracted from the sample.   The extent  of matrix  interferences
       will vary  considerably from source to  source.

4 . 0    SAFETY

4.1    The  toxicity or carcinogenicity of each reagent used in  this method has
       not  been precisely determined;  however, each chemical should be treated
       as a potential  health hazard.   Exposure to these  reagents  should be
       reduced to the  lowest possible  level.   The laboratory is  responsible for
       maintaining a current awareness file of OSHA regulation  regarding  the
       safe handling of the chemicals  specified in this  method.   A reference
       file of data handling sheets should be made available to all personnel
       involved in these analyses.   Specifically,  concentrated  sulfuric acid
       presents some hazards and is moderately toxic and extremely irritating

                                    D-72/SVOA                            OLM04.2

-------
                                          Exhibit D Sernivolatiles -- Appendix A
                                     Screening of Semivolatile Organic Extracts

      to the skin and mucous membranes.  Use these reagents in fume hoods
      whenever possible, and if eye or skin contact occurs, flush with large
      volumes of water.

4.2   Always wear safety glasses or a shield for eye protection, protective
      clothing, and observe proper mixing when working with these reagents.

5.0   EQUIPMENT AND SUPPLIES

      Brand names, suppliers, catalog and part numbers, are for illustrative
      purposes only.  No endorsement is implied.  Equivalent performance may
      be achieved using equipment and supplies other than chose specified
      here, but demonstration of equivalent performance meeting the
      requirements of this SOW is the responsibility of the Contractor.

5.1   Glassware

5.1.1    Continuous Liquid-Liquid Extractors - equipped with Teflon or glass
         connecting joints and stopcocks requiring no lubrications  (Hershberg-
         Wolf Extractor, Ace Glass Company, Vineland, NJ.  P/N 6841 10 or
         equivalent).

5.1.2    Beakers  - 400 mL

5.1.3    Syringes - 0.5 mL

5.1.4    Glass Scintillation Vials - at least 20 mL, with screw cap and Teflon
         or aluminum  foil liner.

5.1.5    Vials and Caps - 2 mL capacity for GC auto sampler.

5.1.6    Disposable Pipets - Pasteur, 1 mL.

5.1.7    Drying Column  - 19 mm ID chromatographic column with coarse  frit
          (substitution of a small pad of Pyrex glass wool for the frit will
         prevent  cross contamination of sample extracts).

5.2   Kuderna-Danish  (K-D) Apparatus

5.2.1    Concentrator Tubes - 15 mL and 10 mL graduated  (Kontes K-570050-1025
         and 570040-1025 or equivalent).  Calibrations must be checked at  the
         volumes  employed in the test.  Ground-glass stoppers are used to
         prevent  evaporation of extracts.

5.2.2    Evaporative  Flasks - 500 mL  (Kontes K-570001-0500 or equivalent).
         Attach to concentrator tube with springs.

5.2.3    Snyder Column  - Two-ball Micro  (Kontes  K-569001-0219 or equivalent).

5.3   Spatula - stainless steel or Teflon

5.4   Balances - Analytical, capable of accurately weighing ± 0.0001  g, and
      one  capable of weighing 100 g to ± 0.01 g.  The balances must  be
      calibrated with class S weights or known reference weights once per  each
      12-hour work shift.  The balances must be-  calibrated with class S

                                   D-73/SVOA                             OLM04.2

-------
Exhibit D Semivolatiles -- Appendix A
Screening of Semivolatile Organic Extracts

      weights at a minimum of once per month.  The balances must also be
      annually checked by a certified technician.

5.5   Ultrasonic Cell Disrupters - Heat Systems, Ultrasonic Inc.,  Model W-385
      Sonicator  (475 watt with pulsing capability, No.  200, M inch tapped
      disruptor horn, No. 419 1/8 inch standard tapered Microtip probe, and
      No. 305, 3/4 inch tapered high gain  "Q" disrupter horn, or No. 208 3/4
      inch standard Solid disrupter horn), or equivalent devices with a
      minimum of 375 watt output capability.

      NOTE: In order to ensure that sufficient energy is transferred to the
      sample during extraction, the Microtip probe or horn  must be replaced if
      the tip begins to erode.  Erosion of the tip is evidenced by a rough
      surface.

5.6   Sonabox Acoustic Enclosure - recommended with above disruptors for-
      decreasing cavitation sound.

5.7   Vacuum Filtration Apparatus

5.7.1    Buchner Funnel

5.7.2    Filter paper  - Whatman No. 41 or  equivalent.

5.8   Pyrex Glass  Wool  -  rinsed with methylene chloride.

5.9   Silicon Carbide Boiling Chips - approximately 10/40 mesh.  Heat  to
      400  °C  for 30 minutes or Soxhlet extract with methylene chloride.

5.10  Water Bath  - heated, with concentric ring  cover,  capable .of temperature
      control  (±2  °C) .   The bath  should  be used in a  hood.

5.11  Nitrogen Evaporation Device  - equipped with a water bath that can be
      maintained at  35-40 °C.  (N-Evap by Organomation Associates,  Inc.,
      South Berlin,  MA,  or equivalent).   To prevent the release  of  solvent
      fumes into the  laboratory,  the nitrogen  evaporation device must  be  used
      in a hood.

5.12  Gas  Chromatograph/Flame lonization Detector (GC/FID System)

5.12.1   Gas  Chromatograph -  an  analytical system complete  with  a temperature
         programmable gas chromatograph  and all required accessories  including
          syringes, analytical columns, and gases.  The  injection port  must  be
          designed for on-column  injection when using  packed columns and  for
          splitless injection  when using  capillary columns.

5.12.2    Gas  Chromatograph Column -  30 m (or longer)  X 0.32 mm,  1 micron  film
          thickness,  silicone  coated,  fused silica capillary column  (J&W
          scientific DB-5 or Equivalent).  Note that this is minimum
          requirement for column  length.   Longer columns.may be used.

5.12.3    Flame ionization Detector
                                    D-74/SVOA                            OLM04.2

-------
6.0
                                          Exhibit D Semivolatiles -- Appendix A
                                     Screening of Semivolatile Organic Extracts
      REAGENTS AND STANDARDS
6.1   Reagents

6.1.1    Reagent water - defined as water in which an interferant is not
         observed at or above the CRQL for each analyte of interest.

6.1.2    Sodium Thiosulfate - (ACS) granular.

6.1.3    Sulfuric Acid Solution  (1+1) - slowly add 50 mL of concentrated H  2SO4
         (sp. gr. 1.84) to 50 mL of reagent water.

6.1.4    Acetone, methanol, methylene chloride - pesticide residue analysis
         grade or equivalent.

6.1.5    Sodium Sulfate - anhydrous powdered reagent grade.  Purify by heating
         at 400  °C for four hours in a shallow tray.  Cool in a desiccator and
         store in a glass bottle  (Baker anhydrous powder, catalog #73898 or
         equivalent).

6.2   Standards

6.2.1    Introduction

         The Contractor must provide all the standard solutions to be used
         with this contract.  These standards may be used only after they have
         been certified according to the procedure in Exhibit E.  The
         Contractor must be able to verify that the standards are certified.
         Manufacturer's certificates of analysis must be retained by the
         Contractor and presented upon request.

6.2.2    Stock Standard Solutions

6.2.2.1     Prepare or purchase  a stock standard solution containing phenol,
            phenanthrene, and di-n-octyl phthalate at concentrations of 1
            ug/uL.  Prepare stock standard solutions by accurately weighing
            about 0.01 g of pure material.  Dissolve the material in pesticide
            quality methylene chloride and dilute to volume in a 10 mL
            volumetric flask.  Larger volumes may be used at the convenience
            of the analyst.  If  compound purity is assayed at 97.0 percent or
            greater, the weight  may.be used without correction to calculate
            the  concentration of  the stock standard.  If the compound purity
            is assayed to be less than 97.0 percent, the weight must be
            corrected when calculating the concentration  (see Exhibit E,
            Analytical Standards Requirements).  Commercially prepared stock
            standards may be used at any concentration if they are certified
            by the manufacturer  or by an independent source  (see Exhibit E).

6.2.2.2     -Transfer the stock standard solutions into Teflon-sealed screw-cap
            bottles.  Stock standard solutions should be checked frequently
            for  signs of degradation or evaporation, especially just prior  to
            preparing calibration standards from them.  Stock standard
            solutions must be replaced after six months or sooner if
            comparison with quality control check samples indicates a problem.
                                    D-75/SVOA
OLM04.2

-------
Exhibit D Semivolatiles -- Appendix A
Screening of Semivolatile Organic Extracts

6.2.3    Working Standard Solutions

6.2.3.1     Surrogate Standard Spiking Solution

            Prepare a surrogate standard  spiking solution that contains
            nitrobenzene-d5, terphenyl-d14/  2-fluorobiphenylt  and 1,2-
            dichlorobenzene-d4 at a  concentration of 100 ug/mL;  phenol-d 5,
            2,4,6-tribromophenol, 2-fluorophenol, and  2-chlorophenol-d  4 at a
            concentration of 150 ug/mL.   Surrogate  standards  are added  to  all
            samples and calibration  solutions.  Additional surrogates may  be
            added  at the laboratory's discretion.

6.2.3.2     GC Calibration  Standard

            Prepare a working standard mixture of phenol, phenanthrene, and
            di-n-octylphthalate.  The concentration must be such that the
            volume injected equals 50 ng  of  each compound.

6.2.4    Storage of Standards

6.2.4.1     Store  the  stock standard solutions at 4 °C  (± 2°C) in Teflon-lined
            screw-cap  amber bottles  and protect from light.

6.2.4.2     Store  the  working standards at  4 °C (±  2°C) in Teflon-sealed
            containers.  The solutions should  be checked frequently for
            stability.  These solutions must be replaced after  six  months  or
            sooner if  comparison with quality  control  check samples indicates
            a problem.

6.2.4.3     Samples,  sample extracts, and standards must be stored  separately.

7.0    QUALITY CONTROL

7.1    Method  Blank

7.1.1     Summary

         A method  blank is  a volume  of a  clean reference matrix (reagent water
          for  water samples, or purified sodium sulfate for soil/sediment
          samples)  that is  carried  through the  entire analytical procedure.
          The  volume  or weight of the reference matrix  must be approximately
          equal to  the volume or weight of samples  associated  with the blank.
          The  purpose  of a  method blank  is to determine the levels of
          contamination associated  with the  processing  and  analysis  of samples.

7.1.2     Frequency

          One  method  blank must be  extracted and  analyzed  on  each GC/FID system
          used to  screen samples  for  the  following,  whichever  is most frequent.
                                    D-76/SVOA         .                   OLM04.2

-------
                                          Exhibit  D Semivolatiles --  Appendix A
                                     Screening  of  Semivolatile  Organic  Extracts
            Each SDG, or
         »  Each 20 samples in a n SDG,  excluding matrix spikes/matrix spike
            duplicates,  that are  of  similar matrix  (water, soil, or sediment)
            or similar concentration  (soil only), or

         •  Whenever samples are  extracted by the same procedure.

7.1.3    Procedure

         For semivolatile analyses, a method blank for water samples consists
         of a 1 L volume of reagent water spiked with 0.5 mL of the surrogate
         spiking solution.  For medium and low level soil/sediment samples,
         method blanks consist of 1 g and 30 g of sodium sulfate spiked with
         0.5 mL of surrogate -spiking solution, respectively.  Extract and
         concentrate method blanks at the same time as the samples associated
         with the blanks according to Section 12.1.  Analyze the method blank
         according to Section 12.1.                    •     '

8.0   CALIBRATION AND STANDARDIZATION

8.1   GC/FID Operating Condition

      Suggested GC operating conditions are as follows:

      •  Initial column  temperature hold - 20  °C for 4 minutes.

      •  Column temperature program - 20-280  °C at 8 "C/minute.

      •  Final column temperature hold - 280  °C for 8 minutes.

      •  Injector - Grob-type, splitless.

      •  Sample volume - 1-2 uL.

      •  Carrier gas - Helium at  30 cm/sec.

8.2   GC Calibration

8.2.1    Summary

         Prior to sample analysis, each GC/FID system must be initially
         calibrated at one concentration level to determine instrument
         sensitivity.

8.2.2    Frequency

         Each GC/FID system must  be calibrated at the beginning of each 12-
         hour shift.

8.2.3    Procedure

         Inject 1-2 uL of the GC  calibration standard prepared in Section
         6.2.3.2.  The volume injected must equal 50 ng of each of the
         calibration compounds.

                                   D-77/SVOA                            OLM04.2

-------
Exhibit D Semivolatiles -- Appendix A
Screening of Semivolatile Organic Extracts

8.2.4    Technical Acceptance Criteria

8.2.4.1      The GC must be  standardized  for  half-scale response  from  50 ng of
             phenanthrene.

8.2.4.2      The GC must adequately  separate  phenol  from the  solvent front.

8.2.4.3      A minimum  of  quarter-scale response  for 50 ng of di-n-
             octylphthalate  must be  exhibited.

8.2.5    Corrective Action

8.2.5.1      If  the technical  acceptance  criteria are not met,  recalibrate the
             GC  instrument.  It may  be necessary  to  change the column  or take
             other corrective  actions to  achieve  the acceptance criteria.

8.2.5.2      GC  calibration  technical acceptance  criteria must be met  before
             any samples  are injected.

9 . 0    PROCEDURE

9.1    Sample Preparation

9.1.1    Low Level Soil/Sediment

9.1.1.1      Decant and discard  any  water layer on a sediment sample.   Mix
             samples  thoroughly,  especially composited samples.  Discard any.
             foreign  objects such as sticks,  leaves, and  rocks.

9.1.1.2      Weigh approximately 30  g of  sample to the nearest 0.1 g into  a
             400 mL beaker and proceed with low level soil/sediment sample
             preparation as  described in  Section 10.1.4.4  of Exhibit D
             (Semivolatiles).

9.1.1.3      Take 5 mL from the  300  mL (approximate) total  extract and
             concentrate to 1  mL following Section 10.2.2.1  or 10.2.2.2 of
             Exhibit  D (Semivolatiles,)  but note that the final volume for
             screening is 1 mL,  not  0.5 mL.

9.1.2     Medium Level Soil/Sediment

9.1.2.1      Decant and discard any water layer on a sediment sample.   Mix
             samples thoroughly,  especially composited samples.  Discard any
             foreign objects such as sticks, leaves, and rocks.

9.1.2.2      Transfer approximately 1 g  (record weight to the nearest 0.1  g)  of
             sample to a 20 mL vial.  Wipe the mouth of the vial with a tissue
             to remove any sample material and proceed with the medium level
             sample preparation procedure described in Section 10.1.4.5 of
             Exhibit D (Semivolatiles).

9.1.2.3      Take 5 mL from the 10 mL total extract and concentrate to 1 mL
             following Section 10.2.2.1 or 10.2.2.2 of Exhibit D
             (Semivolatiles).
                                    D-78/SVOA                             OLM04.2

-------
                                          Exhibit D Semivolatiles -- Appendix A
                                     Screening of Semivolatile Organic Extracts

9.2   GC/FID Analysis

      Inject 1-2 uL of extract.

9.3   Interpretation Of Chromatograms

9.3.1    Soil/Sediment

9.3.1.1     If no sample peaks from the extract  (from low or medium level
            preparation) are detected or all are less than 10.0 percent full
            scale deflection, the sample must be prepared by the low  level
            protocol.

9.3.1.2     Peaks are detected at greater than 10.0 percent full scale
            deflection and less than or equal to full scale deflection.

            •  If the screen is from the medium  level extract, proceed with
               GC/MS analysis of this extract with appropriate dilution if
               necessary.

            •  If the screen is from the low  level extract, discard the
               extract and prepare the samples by medium level method for
               GC/MS analysis.

9.3.1.3     Peaks are detected at greater than full scale deflection.

            •  If the screen is from the medium  level preparation, calculate
               the dilution necessary to reduce  the major peaks  to between
               half  and full scale deflection.   Use this dilution factor to
               dilute the extract.  This dilution is analyzed by GC/MS for
               extractable organics.

            •  If the screen is from the low  level preparation,  discard the
               extract and prepare the samples by the medium  level method for
               GC/MS analysis.
                                   D-79/SVOA                             OLM04.2

-------
       EXHIBIT D
  ANALYTICAL METHODS
FOR PESTICIDES/AROCLORS
        D-l/PEST                            OLM04.2

-------
            Exhibit D - Analytical Methods for Pesticides/Aroclors

                               Table of Contents

Section                                                                    Page

1.0   SCOPE AND APPLICATION  	   4

2 . 0   SUMMARY OF METHOD	   5
      2.1   Water	   5
      2.2   Soil/Sediment  	   5

3 .0   DEFINITIONS	   5

4.0   INTERFERENCES 	   6

5.0   SAFETY	   6

6.0   EQUIPMENT AND SUPPLIES	   7

7.0   REAGENTS AND STANDARDS	13
      7.1   Reagents	  13
            7.1.1 Reagent Water	13
            7.1.2 Sodium Sulfate	13
            7.1.3 Concentrated Sulfuric Acid	13
            7.1.4 Sodium hydroxide solution 	  13
            7.1.5 10 Percent Acetone in Hexane	13
            7.1.6 Methylene Chloride, Hexane, Acetone, Toluene,
                  Iso-octane, and Methanol  	  13
            7.1.7 Mercury    	13
            7.1.8 Copper Powder	13
      7.2   Standards	13
            7.2.1 Introduction	13
            7.2.2 Stock Standard Solutions  	  14
            7.2.3 Secondary Dilution Standards   	  14
            7.2.4 Working Standards	14
            7.2.5 Ampulated Standard Extracts 	  17
      7.3   Storage of Standard Solutions 	  18

8.0   SAMPLE COLLECTION, PRESERVATION, AND STORAGE   	  19
      8.1   Sample Collection and Preservation   	  19
      8.2   Procedure for Sample Storage  	  19
      8.3   Procedure for Sample Extract Storage   	  19
      8.4   Contract Required Holding Times 	  19

9.0   CALIBRATION AND STANDARDIZATION  	  20
      9.1   Gas Chromatograph Operating Conditions   	  20
      9.2  ' Initial Calibration	20
            9.2.1 Summary of Initial Calibration   	  20
            9.2.2 Frequency of Initial Calibration   	  20
            9.2.3 Procedure  for  Initial Calibration  	  21
            9.2.4 Calculations for Initial Calibration   	  21
            9.2.5 Technical Acceptance Criteria for  Initial
                  Calibration	•	25.
            9.2.6 Corrective Action for Initial Calibration  	  27
      9.3   Calibration Verification   .	  .  28
            9.3.1 Summary of Calibration Verification 	  28
            9.3.2 Frequency of Calibration Verification  	  28

                                    D-2/PEST                             OLM04.2

-------
            9.3.3 Procedure for Calibration Verification  	  29
            9.3.4 Calculations for Calibration Verification 	  29
            9.3.5 Technical Acceptance Criteria for
                  Calibration Verification  .	  29
            9.3.6 Corrective Action for Calibration Verification  	  31

10.0  PROCEDURE	 .  .  .  ;	33
      10.1  Sample Preparation	  33
      10.2  GC/EC Analysis	54

11.0  DATA ANALYSIS AND CALCULATIONS	59
      11.1  Qualitative Identification  	 .....  59
            11.1.1      Identification of Target Compounds  	  59
            11.1.2      GC/MS Confirmation of Pesticides and Aroclors  ...  59
      11.2  Calculations	62
            11.2.1      Target Compounds	  	  62
            11.2.2      CRQL Calculation   	  65
            11.2.3      Surrogate Recoveries  	  65
      11.3  Technical Acceptance Criteria for Sample Analysis  	  66
      11.4  Corrective Action for Sample Analysis 	  67

12.0  QUALITY CONTROL	69
      12.1  Blank Analyses	69
            12.1.1      Introduction  	  69
            12.1.2      Method Blanks 	  69
            12.1.3      Sulfur Cleanup Blanks 	  71
            12.1.4      Instrument Blanks  	  73
      12.2  Matrix Spike/Matrix Spike Duplicate  (MS/MSD)   	  74
            12.2.1      Summary of MS/MSD  	  74
            12.2.2      Frequency of MS/MSD Analysis  	  74
            12.2.3      Procedure for Preparing  MS/MSD  	  75
            12.2.4      Calculations for MS/MSD	76
            12.2.5      Technical Acceptance Criteria for MS/MSD   	  76
            12.2.6      Corrective Action  for MS/MSD  	  77

13.0  METHOD PERFORMANCE   	  78

14.0  POLLUTION  PREVENTION  	  78

15.0  WASTE MANAGEMENT	'	78

16.0  REFERENCES	78

17.0  TABLES/DIAGRAMS/FLOWCHARTS   	  79
                                    D-3/PEST                            OLM04.2

-------
  Exhibit D Pesticides/Aroclors -- Section 1
  Scope and Application

  1.0   SCOPE AND APPLICATION

  1.1   In 1978, EPA Headquarters and Regional representatives designed
        analytical methods for the analysis of chlorinated pesticides and
        Aroclors in hazardous waste samples.  These methods were based on EPA
        Method- 608, Organochlorine Pesticides and PCBs.  In 1980, these methods
        were adopted for use in the Contract Laboratory Program  (CLP).   As the
        requirements of CERCLA evolved, the CLP methods, as well as their
        precedent EPA 600 Series methods,  established the basis for other EPA
        methods to perform the analysis of chlorinated pesticides and Aroclors
        in hazardous waste samples (i.e.,  SW-846) .   The following CLP method has
        continuously improved to incorporate technological advancements
        promulgated by EPA, and has continued to set the standard for the
        preparation, extraction, isolation, identification, and reporting of
        chlorinated pesticides and Aroclors at hazardous waste sites.

|  1.2   The analytical method that follows is designed to analyze water,
        sediment, and soil from hazardous waste sites to determine the presence
        and concentration of the chlorinated pesticides and Aroclors found in
        the Target Compound List  (TCL) in Exhibit C.  The method can be used for
        determining analyte concentrations in the range from the contract
        required quantitation limits  (CRQL) to one million times the CRQL in
|        these matrices when appropriate dilutions are made.  The method includes
        sample extraction, extract cleanup techniques and GC/EC analytical
        methods for pesticides and Aroclors.

|  1.3   This analytical method provides the use of SW-846 Methods 3541  (Revision
|        0, September 1994) and 3545  (Revision 0, December 1996) for the
|        extraction of soil/sediment samples.  However, prior to using either one
|        of these alternate extraction procedures, the Contractor must first
|        demonstrate that these procedures are equivalent to the existing
|        procedures, and obtain approval for use of these alternate extraction
|        procedures from the EPA CLP National Program Manager.  The process for
|        determining and documenting equivalency can be found in Exhibit E
|        Section 6.0.

|  1.4   Problems, including resolution difficulties, have been associated with
        the following pairs of compounds using this method.

        •    On  a DB-608 or equivalent column, DDE and dieldrin;  methoxychlor and
             endrin ketone;  and endosulfan I and gamma-Chiordane.

        •    On  a DB-1701  or equivalent column,  endosulfan  I  and  gamma-Chlordane,
             and methoxychlor and  endosulfan sulfate.

|  1.5   There are two isomers of heptachlor epoxide, the endo isomer (isomer A)
        and the exo isomer  (isomer B).  The two isomers are separable using
        current GC capillary'columns.  Only the exo isomer (isomer B) is of
        environmental significance.  This is the isomer that must be used as an
        analytical standard, identified and quantitated in sample analysis, and
        reported on appropriate forms as heptachlor epoxide.
                                     D-4/PEST                             OLM04.2

-------
                               Exhibit D Pesticides/Aroclors -- Sections 2 & 3
                                                 Summary of Method/Definitions

2.0   SUMMARY OF METHOD

2.1   Water

      Continuous liquid-liquid or separator/ funnel extraction procedures are
      employed for aqueous samples.  AIL volume of sample is spiked with the
      surrogate solution and extracted with methylene chloride using a
      separatory funnel or a continuous extractor.  The methylene chloride
      extract is dried with anhydrous sodium sulfate, concentrated, and
      cleaned up by GPC  (GPC is required when higher molecular weight
      compounds are present that interfere with the analyses of target
      compounds; GPC is optional for all other circumstances).  The extract is
      then solvent exchanged into hexane, cleaned up by Florisil cartridges,
      and the final volume adjusted to 1 mL or 2 mL.  The extract is analyzed
      using a dual column wide-bore capillary Gas Chromatography/Electron
      Capture (GC/EC) technique.

2.2   Soil/Sediment

      A 30 g aliquot of sample is spiked with the surrogate and then mixed
      with anhydrous sodium sulfate and extracted with a 1:1 acetone/methylene
      chloride solvent mixture by sonication.  The extract is filtered,
      concentrated and solvent-exchanged into methylene chloride.  The
      methylene chloride extract is then cleaned up by GPC  (mandatory).
      solvent-exchanged into hexane, cleaned up by Florisil cartridge, and
      adjusted to a final volume of 1 mL or 2 mL.  The extract is analyzed
      using a dual column wide-bore capillary Gas Chromatography/Electron
      Capture (GC/EC) technique.

3.0   DEFINITIONS

      See Exhibit G for a complete list of definitions.
                                    D-5/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Sections 4 & 5
Interferences/Safety

4.0   INTERFERENCES

4.1   Method interferences may be caused by contaminants in solvents,.
      reagents, glassware, and other sample processing hardware.  These
      contaminants lead to discrete artifacts or to elevated baselines in gas
      chromatograms.   Routinely, all of these materials must be demonstrated
      to be free from interferences under the conditions of the analysis by
      running reagent blank.  Interferences caused by phthalate esters can
      pose a major problem in pesticide analysis.  Common flexible plastics
      contain-varying amounts of phthalates which are easily extracted during
      laboratory operations, so cross-contamination of glassware frequently
      occurs when plastics are handled.  Interferences from phthalates can
      best be minimized by avoiding the use of such plastics in the
      laboratory.

4.2   Matrix interferences may be caused by contaminants that are co-extracted
      from the sample.  The extent of matrix interferences will vary
      considerably from source to source, depending upon the nature of the
      site being sampled.  The cleanup procedures must be used to remove such
      interferences in order to achieve the contract required quantitation
      limits.

5.0   SAFETY

5.1   The toxicity or carcinogenicity of each reagent used in this method has
      not been precisely determined; however, each chemical should be treated
      as a potential health hazard.  Exposure to these reagents should be
      reduced to the lowest possible level.  The laboratory is responsible for
      maintaining a current awareness file of OSHA regulations regarding the
      safe handling of the chemicals specified in this method.  A reference
      file of data handling sheets should be made available to all personnel
      involved in these analyses.  Specifically, concentrated sulfuric acid
      and the 10 N sodium hydroxide solution are moderately toxic and
      extremely irritating to skin and mucous membranes.  Use these reagents
      in a fume hood whenever possible and if eye or skin contact occurs,
      flush with large volumes of water.  Always wear safety glasses or a
      shield for eye protection, protective clothing, and observe proper
      mixing when working with these reagents.

5.2   The following analytes covered by this method have been tentatively
      classified as known or suspected human or mammalian carcinogens:  4,4'-
      DDT, 4,4'-DDD,  the BHCs, and the-Aroclors.  Primary standards of these
      toxic compounds should be prepared in a hood.  A NIOSH/MESA approved
      toxic gas respirator should be worn when the analyst handles high
      concentrations of these toxic compounds.
                                  ' D-6/PEST                             OLM04.2

-------
                                    Exhibit D Pesticides/Aroclors -- Section 6
                                                        Equipment and Supplies
6.0   EQUIPMENT AND SUPPLIES
      Brand names, suppliers, and part numbers are for illustrative purposes
      only.  No endorsement is implied.  Equivalent performance may be
      achieved using equipment and supplies other than those specified here,
      but demonstration of equivalent performance meeting the requirements of
      this Statement of Work is the responsibility of the Contractor.  The
      Contractor must document in its Narrative when it uses equipment and
      supplies other than those specified here.

6.1   Glassware

6.1.1    Continuous Liquid-Liquid Extractors - equipped with Teflon or glass
         connecting joints and stopcocks requiring no lubrication  (Hershberg-
         Wolf extractor, Ace Glass Company, Vineland, NJ P/N 6841-10, or
         equivalent) or Hydrophobic Membrane-based Extractor (Accelerated One
         Step™  Extractor,  Corning series 3195,  or equivalent).

6.1.2    Separatory Funnels  - 2 L with Teflon stopcock.

6.1.3    Beakers.- 400 inL.

6.1.4    Erlenmeyer Flasks - 250 mL.

6.1.5    Syringes - 10 mL with Luerlok fitting, 1 mL or 2 mL.

6.1.6    Vials and Caps - 20 mL and 10 mL  (optional) with screw cap and Teflon
         or aluminum foil liner, 2 mL capacity for GC auto sampler.

6.1.7    Pipets  - glass volumetric 1 mL  or 2 mL.

6.1.8    Centrifuge Tube - 12 to 15 mL with 19 mm ground glass joint
          (optional).

6.1.9    Graduated Cylinder  - 1 L capacity.

6.1.10   Drying  Column  - chromatographic column approximately 400  mm  long x 19
         mm ID,  with coarse  frit.   (Substitution of a small pad of disposable
         Pyrex glass wool for the frit will help prevent cross-contamination
         of sample extracts).

6.1.11   Volumetric Flasks - 10 mL and 1 or 2 mL.

6.1.12   Bottle  or Test Tube - 20 mL with  Teflon-lined screw cap  for  sulfur
         removal and a  glass bottle  -  1  L  volume,  for use in preparation of
         Bio  Beads  for  packing into a  column.

6.1.13   Powder  Funnels - 10 cm diameter,  for  filtration/drying.

6.1.14   Buchner Funnels  - 9 cm diameter,  for  filtration.

6.2   Kuderna-Danish  (K-D) Apparatus.

6.2.1    Concentrator Tubes  - 10 mL, graduated (Kontes K-570040-1029, or
         equivalent).
                                    D-7/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 6
Equipment and Supplies

6.2.2    Evaporative Flasks - 500 mL (Kontes K-470001-0500, or equivalent).

6.2.3    Snyder Column - three-ball macro  (Kontes K-503000-0121, or
         equivalent).

6.2.4    Snyder Column - two-ball micro  (Kontes K-569001 -0219, or
         equivalent).

6.3   Vacuum System  for Eluting Multiple Cleanup Cartridges.

6.3.1    Vac Elute Manifold - Analytichem  International, J.T. Baker, or
         Supelco  (or equivalent).  The manifold design must ensure that there
         is no contact between plastics containing phthalates and sample
         extracts.

6.3.2    Vacuum Trap - made- from a 500 mL  sidearm flask fitted with a one-hole
         stopper and glass tubing.

6.3.3    Vacuum Pressure Gauge.

6.3.4    Rack for holding 10 mL volumetric flasks in the manifold.

         NOTE:  Other types of equivalent  systems, such as an automated system
         using syringe pressure, are considered to be acceptable for elution
         of florisil cartridges, as long as all QC-and sample technical
         acceptance  criteria are met.

6.4   pH Paper - wide range  (Hydrion Papers, Micro-essential Laboratory,
      Brooklyn, NY,  or equivalent).

6.5   Spatula - stainless steel or Teflon.

6.6   Centrifuge  - table top  (optional).

6.7   Balances -  top loading, capable of weighing accurately to ± 0.01 g,
      analytical, capable of weighing accurately to ± 0.0001 g.  The balances
      must be calibrated with class S weights or known reference weights once
      per each 12-hour work shift.  The balances must be calibrated with class
      S weights at a minimum of once per month.  The balances must also be
      annually checked by a certified technician.

6.8   Ultrasonic  Cell Disrupter - Heat Systems, Ultrasonics, Inc., Model W-385
       (475 watt with pulsing capability, No. 207 3/4-inch tapered disrupter
      horn) or equivalent device with a minimum 375 watt output capability.
      NOTE:  In order to ensure that sufficient energy is transferred to the
      sample during  extraction, the horn must be replaced if the tip begins to
      erode.  Erosion of the tip is evidenced by a rough surface.

6.9   Sonabox Acoustic Enclosure  (or equivalent) - for use with disrupter to
      decrease noise level.

6.10  Filter Paper - No. 41 Whatmann (or equivalent), 9 cm circles  (optional).

6.11  Pyrex Glass Wool - rinsed with methylene chloride and dried before use.

6.12  Boiling chips.

                                   D-8/PEST                             OLM04.2

-------
                                    Exhibit D Pesticides/Aroclors -- Section 6
                                                        Equipment and Supplies

6.12.1   Silicon carbide boiling chips - approximately 10 to 40 mesh.  Heat
         the chips to 400 °C for 30 minutes or solvent rinse before use.

6.12.2   Teflon boiling chips (optional) - solvent rinse the chips before use.

6.13  Water Bath - heated, with concentric ring cover, capable of temperature
      control.  NOTE:  The water bath should be used in a hood.

6.14  GPC Cleanup System

6.14.1   Gel Permeation Chromatography System - GPC Autoprep Model 1002 A or
         B, Analytical Biochemical Laboratories, Inc., or equivalent.  Systems
         that perform satisfactorily have been assembled from the following
         components:  an HPLC pump, an auto sampler or a valving system with
         sample loops, and a fraction collector.  All systems, whether
         automated or manual,, must meet the calibration requirements of
         Section 10.1.8.1.  NOTE:  GPC cleanup is required for extracts for
         all soils/sediments and for water extracts containing higher
         molecular weight contaminants that interfere with the analyses of the
         target compounds.

6.14.1.1    Chromatographic column - 700 mm x 25 mm ID glass column.  Flow is
            upward.  To simplify switching from the UV detector during
            calibration to the GPC collection device during extract cleanup,
            an optional double 3-way valve  (Rheodyne Type 50 Teflon Rotary
            Valve #10-262 or equivalent) may be attached to that the column
            exit flow can be shunted either to the UV flow-through cell or to
            the GPC  collection device.

6.14.1.2    Guard column  (optional) - 5 cm, with appropriate fittings to
            connect  to the inlet side of the.analytical column  (Supelco 5-8319
            or equivalent).

6.14.1.3    Bio Beads  (SX-3) - 200 to 400 mesh, 70 g  (Bio-Rad Laboratories,
            Richmond, CA, Catalog  152-2750, or equivalent).  An additional 5 g
            of Bio Beads  is required if the optional guard column  is employed.
            The quality of Bio Beads may vary from lot to lot because of
            excessive fines in some lots.   In addition to fines having a
            detrimental effect on  Chromatography, they can also pass through
            the column screens and damage  the valve.

6.14.1.4    Ultraviolet detector - fixed wavelength  (254 nm) with  a  semi-prep
            flow-through  cell.

6.14.1.5    Strip chart recorder - recording  integrator or  laboratory data
            system.

6.14.1.6    Syringe  filter  assembly, disposable - Bio-Rad  "Prep Disc" sample
            filter assembly #343-0005,  25  mm, and 5 micron  filter  discs or
            equivalent.   Note:  Some  instrument manufacturer's recommend a
            smaller  micron  filter  disc.  Consult your  instrument operation
            manual to determine the proper size filter disc  to use in your
            system.   Check  each batch for  contaminants.  Rinse each filter
            assembly (prior to use) with methylene chloride  if necessary.
                                    D-9/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 6
Equipment and Supplies

6.15  Florisil - 500 mg or 1 g cartridges with stainless steel or Teflon
      frits, (catalog No. 694-313, Analytichem, 24201 Frampton Ave., Harbor
      City, CA, or equivalent).

6.16  Nitrogen Evaporation Device - equipped with a heated bath that can be
      maintained at 35 to 40 °C (N-Evap by Organomation Associates,  Inc.,
      South Berlin, MA, or equivalent).

6.17  Oven - drying.

6.18  Desiccator.

6.19  Crucibles - porcelain crucibles or aluminum weighing pans.

6.20  pH Meter - with a combination glass electrode.  Calibrate according to
      manufacturer's instructions.  pH meter must be calibrated prior to each
      use.

6.21  Magnetic Stirrer Motor - Model PC 353, Corning Co., Corning, NY, or
      equivalent.

6.22  Magnetic Stirrer Bar - Teflon coated, at least 4 cm long.

&.23  Gas Chromatograph/Electron Capture Detector  (GC/EC) System.

6.23.1   Gas Chromatograph - must adequately regulate temperature in order to
         give a reproducible temperature program and have a flow controller
         that maintains a constant column flow rate throughout temperature
         program operations.  The system must be suitable for split less
         injection and have all required accessories including syringes,
         analytical columns, and gases..

6.23.2   Gas chromatographs that are available from some manufacturers may
         have difficulty in meeting certain method QC requirements because of
         endrin and DDT breakdown in the injector.  This problem can be
         minimized by operating the injector at 200 - 205 °C,  using a Pvrex
         (not quartz) methyl silicone deactivated injector liner, and
         deactivating the metal parts in the injector with dichlorodimethyl
         silane.  In some cases, using a 0.25 inch packed column injector
         converted for use with 0.53 mm capillary columns works better than a
         Grob-type injector.  If a Grob-type injector is used, a 4 mm liner  •
         may be required to meet breakdown criteria.

6.23.3   Gas Chromatograph Columns - two wide-bore  (0.53 mm ID) fused silica
         GC columns are required.  A separate detector is required for each
         column.  The specified analytical columns are a 30 m x 0.53 mm ID,
         1.0 urn film thickness, DB-1701  (J&W Scientific); SPB 1701  (Supelco);
         AT 1701  (Alltech); RTX-1701  (Restek); CP-Sil 19CB (Chrompack); 007-
         1701  (Quadrex); BP-10  (SGE); or equivalent, and a 30 m x 0.53 mm ID,
         0.5 to 1.0 urn film thickness DB-608 (J&W Scientific); HP-608 (Hewlett
         Packard); SPB-608  (Supelco); 007-608  (Quadrex); BP-608  (SGE); CP-Sil
         8CB (Chrompack); or equivalent.  NOTE:  The column length stated
         above is the minimum requirement.  Longer columns that meet
         resolution and calibration requirements may be used.  A description
         of the GC columns used for analysis shall be provided in the SDG
         Narrative.

                                   D-10/PEST                            OLM04.2

-------
                                    Exhibit D Pesticides/Aroclors -- Section 6
                                                        Equipment and Supplies

6.23.3.1    PACKED COLUMNS CANNOT BE USED
 . ^
6.23.3.2    A capillary column is considered equivalent if:

            •  The column does not introduce contaminants which interfere with
               identification and quantitation of the compounds listed in
               Exhibit C  (Pesticides)

            •  The analytical results generated using the column meet the
               initial calibration and calibration verification technical
               acceptance criteria listed in the SOW and the CRQLs listed in
               Exhibit C  (Pesticides).

            •  The column can accept at least 16 times the low point standard
               for individual standard mixtures A and B for each compound
               listed in Exhibit C  (Pesticides) without becoming overloaded.

            •  The column pair chosen must have dissimilar phases/chemical
               properties in order to separate the compounds of interest in
               different RT order.

6.23.3.3    Although the  instructions included in the SOW are for wide bore
            capillary columns, narrower bore capillary columns may be
            .evaluated for use.                    .

6.23.3.4 .   As. applicable, follow the manufacturer's instructions for use of
            • its product.

6.23.3.5    The Contractor must maintain documentation that the column met the
            criteria in Section 6.23.3.2.  The minimum documentation is as
            follows:

6.23.3.5.1     Manufacturer provided information concerning the performance
               characteristics of the column;

6.23.3.5.2     Chromatograms  and data system reports generated on the GC/ECD
               and used for CLP analyses:

               •  From  instrument blanks which demonstrate that there are no
                  contaminants which interfere with the pesticide analysis
                  when using  the alternate columns;

               •  For initial calibration  standards analyzed using  the column;

             .. •  For calibration verification standards analyzed using the
                  alternate column.

6.23.3.5.3     Based on the Contractor generated data  des.cribed  in  Section
               6.23.3.5.2,  the Contractor  must complete a  written review,
               aligned'by the Laboratory Manager certifying that:

               •  The column  performance is  comparable to  the  required column
                  performance in  its ability to produce initial  calibration
                  and calibration verifications which  meet the technical
                  acceptance  criteria in Sections 9.2.5 and 9.3.5.
                                   D-11/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 6
Equipment and Supplies

               •  The low point initial calibration standard analyses have
                  adequate sensitivity to meet the pesticide CRQLs.

               •  The high point initial calibration standard analyses were
                  not overloaded.

               •  The column does not introduce contaminants which interfere
                  with identification and quantitation of compounds listed in
                  Exhibit C  (Pesticides).                             '

6.23.3.5.4     The documentation must be made available to the Agency during
               on-site laboratory evaluations or sent to the Agency upon
               request of the Technical Project Officer or the Administrative
               Project Officer.

6.23.3.6    Columns are mounted in a 0.25-inch injector ports, by using glass
            adapters available from a variety of commercial sources  (J&W
            Scientific, Supelco, Inc., Hewlett-Packard, Varian, Inc., Perkin
            Elmer, or equivalent) .   The two columns may be mounted into a
            single injection port with a tee adapter (Supelco, Inc.,
            Bellefonte, PA, Catalog No. 2-3660, or equivalent).  Use of this
            adapter allows simultaneous injection onto both columns.  The
            laboratory should follow manufacturer's recommendations for
            mounting 0.53 mm capillary columns in injector ports.

6.23.3.7    The carrier gas for routine applications is helium..  Laboratories
            may choose to use hydrogen as a carrier gas, but they must clearly
            identify its use in the SDG Narrative and on all divider pages
            preceding raw chromatographic data in submissions to the Agency.
            Laboratories that choose to use hydrogen are advised.to exercise
            caution in its use.  Use of a hydrogen leak detector is highly
            recommended when hydrogen is used as the carrier gas.  All GC
            carrier gas lines must be constructed from stainless steel or
            copper tubing.  Non-polytetrafluoroethylene  (PTFE) thread sealants
            or flow controllers with rubber components are not to be used.

6.23.4   Electron Capture Detector (BCD) - the linearity of the response of
         the ECD may be greatly dependent on the flow rate of the make-up gas.
         The make-up gas must be P-5, P-10  (argon/methane) or nitrogen
         according to the instrument specification.  Care must be taken to
         maintain stable and appropriate flow of make-up gas to the detector.
         The GC/EC system must be in a room in which the atmosphere has been
         demonstrated to be free of all contaminants which may interfere with
         the analysis.  The instrument must be vented to outside the facility
         or to a trapping system which prevents the release of contaminants
         into the instrument room.

6.23.5   Data System - a data system must be interfaced to the GC/EC.  The
         data system must allow the continuous acquisition of data throughout
         the duration of the chromatographic program and must permit, at the
         minimum, the output of time vs. intensity  (peak height or peak area)
        • data.  Also, the data system must be able to rescale chromatographic
         data in order to report chromatograms meeting the requirements listed
         within this method.
                                   D-12/PEST                            OLM04.2

-------
                                    Exhibit D Pesticides/Aroclors --  Section 7
                                                        Reagents and  Standards

7.0   REAGENTS AND STANDARDS  .

7.1   Reagents

7.1.1    Reagent Water - defined as water in which no interferant is  observed
         at one-half the CRQL of any pesticide/Aroclor when one liter of the
         reagent water is extracted and prepared by using the same workup
         procedure as for a water sample.

7.1.2    Sodium Sulfate - granular-anhydrous reagent grade, heated at 400 °C
         for 4 hours, or at 120 °C for 16 hours,  cooled in a desiccator, and
         stored in a glass bottle.  Each lot must be extracted with hexane and
         analyzed by GC/EC to demonstrate that it is free of interference
         before use.  Baker anhydrous granular, Catalog No. 3375, or
         equivalent.  CAUTION:  An open container of sodium sulfate may become
         contaminated during storage in the laboratory.

7.1.3    Concentrated Sulfuric Acid  (H2SO4)- 18 N.

7.1.4    Sodium Hydroxide Solution  (NaOH)  (10 N)  - carefully dissolve 40 g of
         NaOH in reagent water and dilute the solution to 100 mL.

7.1.5    10 Percent Acetone in Hexane  (v/v) - prepare by adding 10 mL of
         acetone to 90 mL of hexane.  NOTE:  Prepare this mixture accurately
         or the results from the Florisil cartridge cleanup will be adversely
         affected.  Water in the acetone also will adversely affect Florisil
         performance.

7.1.6    Methylene Chloride, Hexane, Acetone, Toluene, Iso-octane, and
         Methanol  (optional) - pesticide quality or equivalent.  It is
         recommended that each lot of solvent used be analyzed to demonstrate
         that it is free of interference before use.  Methylene chloride must
         be certified as acid free or must be tested to demonstrate that it is
         free of hydrochloric acid.  Acidic methylene chloride must be  passed
         through basic alumina and then demonstrated to be  free of
         hydrochloric acid.

7.1.7    Mercury - triple distilled, for sulfur cleanup.

7.1.8    Copper Powder  (optional)  -  fine,  granular  (Mallinckrodt 4649 or
         equivalent).  Copper may be used  instead of mercury for sulfur
         cleanup.  Remove oxides by  treating with dilute nitric acid, rinse
         with distilled water to remove  all traces of acid, rinse with
         acetone,  and dry under a  stream of nitrogen.

7.2   Standards

7.2.1    Introduction

         The  Contractor must provide all standards to be used with this
         contract.   These standards  may  be used only after they have been
         certified according to the  procedure  in Exhibit E.  The Contractor
         must be able to verify that the standards are certified.
         Manufacturer's  certificates of  analysis must be retained by the
         Contractor  and presented  upon request.
                                   .D-13/PEST                            OLM04.2

-------
Exhibit D Festicides/Aroclors -- Section 7
Reagents and Standards

7.2.2    Stock Standard Solutions  (1 ug/uL) - can be prepared from pure
         standard materials or purchased as certified solutions.

7.2.2.1     Prepare stock standard solutions by accurately weighing about
            0.0100 g of pure material.  Dissolve the material in toluene and
            dilute to volume in a 10 mL volumetric flask with toluene or
            acetone.  Larger volumes may be used at the convenience of the
            analyst.

7.2.2.2     When compound purity is assayed to be 97 percent or greater, the
            weight may be used without correction to calculate the
            concentration of the stock solution.  If the compound purity is
            assayed to be less than 97 percent, -the weight must be corrected
            when calculating the concentration of the stock solution.   (See
            Exhibit E Analytical Standards Requirements.)

7.2.2.3     Fresh stock standards must be prepared once every six months, or
            sooner if standards have degraded or concentrated.  Stock
            standards must be checked for signs of degradation or
            concentration just prior to preparing working standards from them.

7.2.3    Secondary Dilution Standards

         Using stock standards, prepare secondary dilution standards in
         acetone that contain the compounds of interest either singly or mixed
         together.  Fresh secondary dilution standards must be prepared once
         every six months, or sooner if standards have degraded or
         concentrated.  Secondary dilution standards must be checked for signs
         of degradation or concentration just prior to preparing working
         standards from them.

7.2.4    Working Standards

7.2.4.1     Surrogate Standard Spiking Solution

            The surrogates, tetrachloro-m-xylene and decachlorobiphenyl, are
            added to all standards, samples, matrix spikes, and blanks.
            Prepare a surrogate spiking solution of 0.2 ug/mL of each of the
            two compounds in acetone.

7.2.4.2     Matrix Spiking Solution

            Prepare a matrix spiking solution in acetone or methanol that
            contains the following pesticides at the concentrations specified:


                   Pesticide                    Concentration ug/mL
                   gamma-BHC (Lindane)            .     0.5
                   4,4'-DDT                           1.0
                   Endrin                             1.0
                   Heptachlor                         0.5
                   Aldrin                             0.5
                   Dieldrin                           1.0
                                   D-14/PEST                            OLM04.2

-------
                                      Exhibit D Pesticides/Aroclors -- Section 7
                                                          Reagents and Standards

|  7.2.4.3     GPC Calibration and Calibration Verification Solutions

|  7.2.4.3.1      Prepare a GPC calibration solution in methylene chloride that
                 contains the following analytes at the minimum concentrations
                 listed below:

                     Analvte                      Concentration mq/mL
                     Corn oil                      .     25.0
                     Bis-2-ethylhexyl phthalate          0.5
                     Methoxychlor                        0.1
                     Perylene                            0.02
                     Sulfur    .                          0.08

  7.2.4.3.2      NOTE:  Sulfur is not very soluble in methylene chloride, but it
                 is soluble  in warm corn oil.  Therefore, one approach is to
                 weigh out the corn oil, warm it, and transfer .the weighed
                 amount of sulfur into the warm corn oil.  Mix it and then
                 transfer into a volumetric  flask with methylene chloride, along
                 with the other calibration  compounds.

|  7.2.4.3.3      GPC Calibration Verification Solution

                 Prepare a GPC calibration verification  solution in methylene
                 chloride  that contains  the  following compounds.  The
                 concentrations listed below are for a 5 mL GPC injection loop.
                 See'section 10.1.8.1.4.3 for compound concentrations if a
                 smaller size loop  is being  used.

                     Compound                     Concentration UQ/mL

                     gamma-BHC  (Lindane)                  0.1
                     Heptachlor                           0.1
                     Aldrin                               0.1
                     4,4 '-DDT                            0.2
                     Endrin                               0.2
                     Dieldrin                            0.2

                  The Aroclor mixture contains  2  ug/mL  each  of Aroclor 1016  and
                  1260  in  methylene  chloride.

  7.2.4.4      Florisil  Cartridge Check  Solution

               Prepare  a solution of 2,4,5-Trichlorophenol in acetone, at a
               concentration of  0.1  ug/mL.
                                      P-15/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 7
Reagents and Standards

7.2.4.5     Resolution Check Mixture

            Prepare a mixture in hexane or iso-octane that contains the
            following pesticides and surrogates at the concentrations listed
            below.

               Compound                   Concentration  (ng/mL)
               gamma-Chlordane                  10.0
               Endosulfan I                     10.0
               4,4'-DDE                         20.0
               Dieldrin                         20.0
               Endosulfan sulfate.               20.0
               Endrin ketone                    20.0
               Methoxychlor                     100.0
               Tetrachloro-m-xylene             20.0
               Decachlorobiphenyl               20.0

7.2.4.6     Performance Evaluation Mixture  (PEM)

            Prepare the PEM in hexane or iso-octane at the concentration
            levels listed below.

               Compound                   Concentration  (ng/mL)
               gamma-BHC                        10.0
               alpha-BHC                        10.0
               4, 4'-DDT                         100.0
               beta-BHC                         10.0
               Endrin                           50.0
               Methoxychlor                     250.0
               Tetrachloro-m-xylene             20.0
               Decachlorobiphenyl               20.0

7.2.4..7     Individual Standard Mixtures A and  B

            The  single component pesticide standards must be prepared in
            hexane or iso-octane at three concentrations for each analyte,
            including the surrogates.  Two separate calibration mixtures, A
            and  B  (listed below), are used to ensure that each peak is
            adequately resolved.  The low point concentration corresponds to
            the  CRQL for each analyte.  The midpoint concentration must be 4
            times  the low point concentration.  The high point concentration
            must be at least 16 times that of the  low point, but a higher
            concentration may be chosen by the  Contractor.  The high point
            concentration defines the upper end of the concentration range for
            which  the calibration is valid.
                                   D-16/PEST                            OLM04.2

-------
                                    Exhibit D  Pesticides/Aroclors  --  Section 7
                                                        Reagents and  Standards

                        Individual Standard Mixture A
                           Low Point Concentration

                  alpha-BHC                                5.0  ng/mL
                  Heptachlor                               5.0  ng/mL
                  gamma-BHC                                5.0  ng/mL
                  Endosulfan  I                             5.0  ng/mL
                  Dieldrin                                10.0  ng/mL
                  Endrin                                  10.0  ng/mL
                  4,4'-DDD                                10.0  ng/mL
                  4,4'-DDT                                10.0  ng/mL
                  Methoxychlor                            50.0  ng/mL
                  Tetrachloro-m-xylene                     5 . 0  ng/mL
                  Decachlorobiphenyl                      10.0  ng/mL
                        Individual Standard Mixture B
                           Low Point Concentration

                   beta-BHC
                   delta-BHC
                   Aldrin
                   Heptachlor Epoxide
                        (exo-epoxy isomer)
                   alpha-Chlordane
                   gamma-Chlordane
                   4,4'-DDE
                   Endosulfan sulfate
                   Endrin  aldehyde
                   Endrin  ketone
                   Endosulfan II
                   Tetrachloro-m-xylene
                   Decachlorobiphenyl
 5.0 ng/mL
 5.0 ng/mL
 5.0 ng/mL
 5.0 ng/mL

 5.0 ng/mL
 5.0 ng/mL
10.0 ng/mL
10.0 ng/mL
10.0 ng/mL
10.0 ng/mL
10.0 ng/mL
 5.0 ng/mL
10.0 ng/mL
            NOTE:  Only the exo-epoxy isomer (isomer B)  of heptachlor epoxide
                   is used as an analytical  standard.

7.2.4.8     Multicomponent Standards

            Toxaphene and Aroclor standards must be prepared individually
            except for Aroclor 1260 and Aroclor 1016 which may be combined in
            one standard mixture.  The calibration standards for the Aroclors
            must be prepared at concentrations of 100 ng/mL, except for
            Aroclor 1221 which must be prepared at 200 ng/mL.  Toxaphene must
            be prepared at 500 ng/mL.  All multicomponent standards must
            contain the surrogates at 20 ng/mL.  The Aroclor and toxaphene
            solutions must be prepared in hexane or iso-octane.

7.2.5    Ampulated Standard Extracts

         Standard solutions purchased from a chemical supply house as
         ampulated extracts in glass vials may be retained until the
         expiration date provided by the manufacturer.  If no manufacturer's
         expiration date is provided, the standard solutions as ampulated
         extracts may be retained and used for 2 years from the preparation
         date.  Standard solutions prepared by the Contractor which are
                                   D-17/PEST
              OLM04.2

-------
  Exhibit D festicides/Aroclors -- Section 7
  Reagents and Standards

           immediately ampulated in glass vials may be retained 2 years from the
           preparation date.  Upon breaking the glass seal, the expiration times
           listed in Sections 7.2.2 through 7.3 will apply.  The Contractor is
           responsible for assuring that the integrity of the standards have not
           degraded (see Section 7.3.5).

  7.3   Storage of Standard Solutions

|  7.3.1    Store the stock and secondary dilution standard solutions at
j     .      4 °C (± 2°C)  in Teflon-lined screw cap amber bottles/vials.   Fresh
           standards should be prepared every six months, or sooner if
           comparison with check standards indicates a problem.

  7.3.2    Store the GPC calibration solution in an amber glass bottle with a
|           Teflon lined screw-cap at 4°C (± 2°C) and protect from light.
           (Refrigeration may cause the corn oil to precipitate.  Before use,
           allow the calibration solution to stand at room temperature until the
           corn oil dissolves.)  Replace the calibration standard solution every
           six months, or more frequently if necessary.

  7.3.3    Store all other working standard solutions in amber glass bottles or
|           vials with Teflon lined screw caps at 4°C (± 2°C) and protect  from
           light.  The standard solution must be checked frequently for
           stability.  Replace all working standard solutions after six months,
           or sooner if comparison with quality control check samples indicates
           a problem, except for the PEM solution which must be prepared weekly.
           CAUTION:  Analysts must allow all standard solutions .'to equilibrate
           to room temperature before use.

  7.3.4    Samples, sample extracts, and standards must be stored separately.

  7.3.5    The Contractor is responsible for maintaining the integrity of
           standard solutions and verifying prior to use.  This means that
           standards must be brought to room temperature prior to use, checked
           for losses, and checked that all components have remained in the
           solution.
                                     D-18/PEST                            OLM04.2

-------
                                      Exhibit D Pesticides/Aroclors -- Section 8
                                    Sample Collection,  Preservation,  and Storage

  8.0   SAMPLE COLLECTION, PRESERVATION,  AND STORAGE

  8.1   Sample Collection and Preservation

  8.1.1    Water samples may be collected in 1 L (or 1 quart)  amber glass
           containers, fitted with screw-caps lined with Teflon.  If amber
           containers are not available,  the samples should be protected from
           light.  Soil samples may be collected in glass containers or closed
           end tubes  (e.g., brass sleeves) in sufficient quantity to perform the
           analysis.  The specific requirements for site sample collection are
           outlined by the Region.

|  8.1.2    All samples must be iced or refrigerated at 4°C (+ 2°C)  from the time
           of collection until extraction.                                     -.

  8.2   Procedure for Sample Storage

  8.2.1    The samples must be protected from light and refrigerated at
           4 °C  (±2 °C)  from the time of  receipt until 60 days after delivery of
           a complete reconciled sample data package to the Agency.  After 60
           days  the samples may be disposed of in a manner that complies with
           all applicable regulations.

  8.2.2    The samples must be stored in an atmosphere demonstrated to be free
           of all potential contaminants.

  8.3   Procedure for Sample Extract Storage

  8.3.1    Sample extracts must be protected from light and stored at
|           4°C  (± 2°C) until 365 days after delivery of a complete reconciled
           data  package to the Agency.

  8.3.2    Sample extracts must be stored in an atmosphere demonstrated  to be
           free  of  all potential contaminants.

  8.3.3    Samples, sample extracts, and  standards must be stored  separately.

  8.4   Contract Required Holding Times

  8.4.1    Extraction of  water samples by separatory funnel procedures must be
           completed  within five days of  the Validated Time of  Sample Receipt
            (VTSR).  Extraction of water samples by continuous liquid-liquid
           extraction procedures must be  started within  five days  of VTSR.
           Extraction of  soil/sediment samples by sonication must  be  completed
           within 10  days of VTSR.

  8.4.2    As part  of the Agency's QA program, the Agency may provide
           Performance Evaluation  (PE) samples as standard extracts which  the
           Contractor is  required to prepare per instructions provided by  the
|           Agency.  PE samples must  be prepared and analyzed  concurrently  with
|            the  samples in the SDG.   The extraction holding times  (five days
           after VTSR for water,  10  days  after VTSR for  soil/sediment) do  not
|           apply for  PE  samples  received  as  standard extracts.

  8.4.3    Analysis of sample extracts must  be completed within 40 days
            following  the  start of extraction.

                                     D-19/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 9
Calibration and Standardization
GC Operating Conditions/Initial Calibration
9.0   CALIBRATION AND STANDARDIZATION

9.1   Gas Chromatograph Operating Conditions

9.1.1    The following are the gas chromatographic analytical conditions.  The
         conditions are recommended unless otherwise noted.-'

            Carrier Gas:     .       Helium  (hydrogen may be used, see Section
                                    6.23.3.7)
            Column Flow:            5 mL/min
            Make-up Gas:            P-5/P-10 or N2 (required)
            Injector Temperature:   > 200 °C (see Section 9.1.4)
            Injection:              On-column
            Injection Volume:       1 or 2 uL  (see Section 9.1.3)
            Injector:               Grob-type, split less
            Initial Temperature:    150 °C
            Initial Hold Time:      M min
            Temperature Ramp:       5 to 6° C/min
            Final Temperature:      275 °C
            Final Hold Time:        Until after decachlorobiphenyl has eluted
                                     (approximately 10 minutes)

9.1.2    Optimize GC conditions for analyte separation and sensitivity.  Once
         optimized, the same GC conditions must be used for the analysis of
         all standards, samples, blanks, and MS/MSD.  The linearity of the BCD
         may be greatly dependent on the flow  rate of the make-up gas.  Care
         must be taken to maintain stable and  appropriate flow of make-up gas
         to the detector.

9.1.3    Manual injections must be 2 uL.  Auto injectors may use 1 uL volumes.
         The same injection volume must be used for all standards, blanks, and
         samples, including MS/MSD.

9.1.4    Cold  (ambient temperature) on-column  injectors that allow injection
         directly onto a 0.53 mm ID column may be used as long as the
         acceptance criteria for resolution, calibration, and analyte
         breakdown are met.

9.2   Initial Calibration

9.2.1    Summary of Initial Calibration

         Prior to the analysis of samples, including MS/MSD and required
         blanks, each GC/EC system must be initially calibrated at a minimum
         of three concentrations to determine  instrument sensitivity and the
         linearity of response utilizing single component target compound and
         surrogate standards.  Multicomponent  target compounds are calibrated
         at a single point.

9.2.2    Frequency of Initial Calibration

         Each GC/ECD system must be initially  calibrated upon award of the
         contract, whenever major instrument maintenance or modification is

                                   D-20/PEST                            OLM04.2

-------
                                    Exhibit D Pesticides/Aroclors -- Section 9
                                               Calibration and Standardization
                                                           Initial Calibration

         performed  (e.g., column replacement or repair, cleaning or
         replacement of BCD, etc.) or if the calibration verification
         technical acceptance criteria have not been met.

9.2.3    Procedure for Initial Calibration

9.2.3.1     Set up the GC/ECD systems as described in Section 9.1.

.9.2.3.2     Prepare the initial calibration standards using the analytes and
            the concentrations specified in Sections 7.-2.4.5 through 7.2.4.8.

9.2.3.3     All standards, samples, MS/MSD, blanks, and extracts must be
            allowed to warm to ambient temperature before analysis.

9.2.3.4     Analyze the initial calibration sequence as given below.  NOTE:
            Steps 16 and  17 are used as part of the calibration verification
            as well  (see  Section 9.3).
                        INITIAL CALIBRATION SEQUENCE
                        1.    Resolution Check
                        2.    Performance Evaluation Mixture
                        3.    Aroclor  1016/1260
                        4.    Aroclor  1221
                        5.    Aroclor  1232
                        6.    Aroclor  1242
                        7.    Aroclor  1248
                        8.    Aroclor  1254
                        9.    Toxaphene
                        10.   Low  Point Standard A
                        11.   Low  Point Standard B
                        12.   Midpoint Standard A
                        13.   Midpoint Standard B
                        14.   High Point Standard A
                        15.   High Point Standard B
                        16.   Instrument Blank
                        17.   Performance Evaluation Mixture

 9.2.4     Calculations  for  Initial  Calibration

 9.2.4.1      During the initial calibration  sequence, absolute  retention  times
             (RT)  are determined for all single component pesticides,  the
             surrogates, and at least three  major peaks  of each multicomponent
             analyte.
                                    D-21/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 9
Calibration and Standardization
Initial Calibration

9.2.4.2  For single component pesticides, an RT is measured in each of three
         calibration standards and the mean RT is calculated as the average of
         the three values.  An RT is measured for the surrogates in each of
         the three analyses of Individual Standard Mixture A during the
         initial calibration and the mean RT is calculated as the average of
         the three values.  Calculate a mean absolute retention time   ("RT)
         for each single component pesticide and surrogate using Equation 1.

            EQ. 1
                                       ^
                                 	   2-i RTi
                                 RT = —	
                                         n
            Where,


             RT    =    Mean absolute retention time of analyte.

             RTi   =    Absolute retention time of analyte.

             n     =    Number of measurements  (3) .

9.2.4.3     A retention time window is calculated  for each single component
            analyte and surrogate and for the major peaks  (3 to 5) of each
            multicomponent analyte by using the list in Table 1.  Windows are
            centered around the mean absolute retention time for the analyte
            established during the initial calibrations.  Analytes are
            identified when peaks are observed in  the RT window for the
            compound on both GC columns.

9.2.4.4     The linearity of the instrument is determined by calculating a
            percent relative standard deviation (%RSD) of the calibration
            factors from a three-point calibration curve for each single
            component pesticide and surrogate.  Either peak area or peak
            height may be used to calculate calibration factors used in the
            %RSD equation.  For example, it is permitted to calculate
            linearity for endrin based on peak area and to calculate linearity
            for aldrin based on peak height.  It is not permitted within a
            %RSD calculation for an analyte to use calibration factors
            calculated from both peak area and peak height.  For example, it
          ,  is not permitted to calculate the calibration factor for the low
            point standard for endrin using peak height and calculate the
            midpoint and high point standard calibration factors for endrin
            using peak area.

9.2.4.5     Calculate the calibration factor for each single component
            pesticide and surrogate over the initial calibration range using
            Equation 2.  The calibration factors for the surrogates are
            calculated from the three analyses of  Individual Standard Mixture
            A only.
                                   D-22/PEST                            OLM04.2

-------
                                    Exhibit D Pesticides/Aroclors -- Section 9
                                               Calibration and Standardization
                                                           Initial Calibration

9.2.4.6     Calculate the mean and the %RSD of the calibration factors for
            each single component pesticide and surrogate over the initial
            calibration range using Equations 3 and 4.
            EQ. 2
   _ Peak area (or height) of the standard
              Mass injected (ng)
            EQ. 3
                                          CF
                                  CF = -i^-
            EQ. 4
                                %RSD =
                  SDcp
                 - —
                  CF
                                           x 100
            Where,
                              K -
                              (n - 1)
             %RSD     = Percent relative standard deviation
                     = Standard deviation of calibration factors
              CF.     = Calibration factor
              CF
= Mean calibration factor
              n      = Total number of values (3)

 9.2.4.7      A calibration factor  is calculated  for each  peak  in a selected set
             of three to five major peaks  for each multicomponent analyte using
             Equation 2.
                                   .D-23/PEST
                                                    OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 9
Calibration and Standardization
Initial Calibration

9.2.4.8     Calculate the percent breakdown of DDT, the percent breakdown of
            endrin, and the combined breakdown of DDT and endrin in the PEM
            using Equations 5, 6, 7, and 8.

            EQ. 5   .   .

                 .  ,.    ,  .  .    Peak area (height) of compound in PEM
            Amount found  (ng)  = 	-—'	—	
                                               CFmp
            Where,
                        The calibration factor for the compound determined
                        from the midpoint standard in the most recent  initial
                        calibration.  NOTE:   If during the  initial
                        calibration, linearity was determined based  on peak
                        area for the compound, then the midpoint  CF  must be
                        based on peak area.   If during the  initial
                        calibration, the linearity for the  compound  was
                        determined based on peak height for the compound, then
                        the midpoint CF must  be based on peak height .
            EQ. 6
                     , ,     „„„    Amount found (ng) (DDD+DDE)
                %Breakdown DDT  = - -^ — - - xlOO
                                 Amount  (ng) of DDT injected
            EQ. 7

                       Amount fo'
                  •in = 	
                                 Amount (ng) of endrin injected
„_   . ,    „ ,  .     Amount found (ng) (endrin aldehyde + enrin ketone) , _.-
%Breakdown Endrin =  	±	£-	xlOO
            EQ. 8

            Combined %Breakdown = %Breakdown DDT + %Breakdown Endrin
9.2.4.9     Calculate the percent difference  for each single  component
            pesticide and surrogate  in the  PEM using Equations  5  and 9.

            EQ.  9                 C    - C
                              %D = -^^	^ x 100
                                      C
                                      nora
            Where,

            %D       =    Percent difference
            Cnom      =    Nominal concentration of each analyte
            ccaic     =    Calculated concentration of each  analyte  from the
                         analyses of  the  standards.
                                   D-24/PEST                            OLM04.2

-------
9.2.4.10
                        Exhibit D Pesticides/Aroclors -- Section 9
                                   Calibration and Standardization
                                               Initial Calibration

Calculate the resolution between the analytes in the Resolution
Check Mixture,, the Performance Evaluation Mixture, and the
midpoint concentration of Individual Standard Mixtures A and B
using Equation 10.
                      Area or Height
                                  Time-
            EQ. 10
                                           V
                             %Resolution = — x 100
                                           H
            Where,
            V  =
            H  =
         Depth of the valley between the two peaks.   The depth of
         the valley is measured along a vertical line from the
         level of the apex of the shorter peak to the floor of the
         valley between the two peaks.
         Height of the shorter of the adjacent peaks.
9.2.5    Technical Acceptance Criteria  for Initial Calibration

         All initial  calibration  technical acceptance criteria apply
         independently  to both GC columns.

9.2.5.1     The  initial calibration  sequence must be analyzed according  to  the
            procedure and in the  order  listed  in Section  9.2.3,  at  the
            concentrations  listed in Sections  7.2.4.5 through 7.2.4.8, and  at
            the  frequency listed  in  Section 9.2.2.  The GC/ECD operating
            conditions  optimized  in  Section 9.1 must be followed.

9.2.5.2     The  resolution  between two  adjacent peaks in  the Resolution  Check
            Mixture must be greater  than or equal to 60.0 percent.

9.2.5.3     All  single  component  pesticide and surrogate  peaks in both runs of
            the  PEM must be greater  than or equal to 90.0 percent resolved  on
            each column.

9.2.5.4     The  absolute retention times of each of the single component
            pesticides  and  surrogates in both  runs of the PEM must  be within
                                   D-25/PEST
                                                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 9
Calibration and Standardization
Initial Calibration

            the retention time windows determined from the three-point initial
            calibration in Section 9.2.4.3.

9.2,5.5     The percent difference of the calculated amount (amount found)  and
            the nominal amount  (amount added) for each of the single component
            pesticides and surrogates in both of the PEM runs of each GC
            column must be greater than or equal to -25.0 AND less than or
            equal to 25.0 percent using Equation 9.

9.2.5.6     The percent breakdown of DDT and endrin in each of the PEM runs
            must be less than or equal to 20.0 percent.  The combined
            breakdown of DDT and endrin must be less than or equal to 30.0
            percent.

9.2.5.7     The %RSD of the calibration factors for each single component
            target compound must be less than or equal to 20.0 percent, except
            alpha-BHC and delta-BHC.  The %RSD of the calibration factors for
            alpha-BHC and delta-BHC must be less than or equal to 25.0
            percent.   The %RSD of the calibration factors for the two
            surrogates must be less than or equal to 30.0 percent.  Up to two
            single component target compounds (but not surrogates) per column
            may exceed the 20.0 percent limit for %RSD (25.0 percent for
            alpha-BHC and delta-BHC), but those compounds must have a %RSD of
            less than or equal to 30.0 percent.

9.2.5.8     The resolution between any two adjacent peaks in the midpoint
            concentrations of Individual Standard Mixtures A and B in the
            initial calibration must be greater than or equal to 90.0 percent.

9.2.5.9     All instrument blanks must meet the technical acceptance criteria
            in Section 12.1.4.4.

9.2.5.10    The identification of single component pesticides by gas
            chromatographic methods is based primarily on retention time data.
            The retention time of the apex of a peak can only be verified from
            an on-scale chromatogram.  The identification of multicomponent
            analytes by gas chromatographic methods is based primarily on
            recognition of patterns of retention times displayed on a
            chromatogram.  Therefore, the following requirements apply to all
            data presented for single component and multicomponent analytes.

            •    The  chromatograms  that  result from the analyses of  the
                 Resolution  Check Mixture,  the PEM,  and Individual Standard
                 Mixtures A  and  B during the  initial calibration sequence must
                 display  the single component analytes present  in  each
                 standard at greater than 10  percent of full scale but  less
                 than 100 percent of full scale.

            •    The  chromatograms  for at least  one of the  three analyses each
                 of Individual Standard  Mixtures A  and B  from the  initial
                 calibration sequence must  display  the single component
                 analytes at greater than 50  percent and  less than 100 percent
                 of full  scale.
                                   D-26/PEST                            OLM04.2

-------
                                    Exhibit D  Pesticides/Aroclors  --  Section  9
                                              Calibration  and Standardization
                                                           Initial Calibration

            •   The chromatograms of the standards for the multicomponent
                analytes analyzed during the initial calibration sequence
                must display the peaks chosen for identification of each
                analyte at greater than 25 percent and less than 100 percent
                of full scale.

            •   For any standard containing alpha-BHC, the baseline of the
                chromatogram must return to below 50 percent of full scale
                before the elution time of alpha-BHC, and return to below 25
                percent of full scale after the elution time of alpha-BHC and
                before the elution time of decachlorobiphenyl.

            •   If a chromatogram is replotted electronically  to meet
                requirements, the scaling factor used must be  displayed on
                the chromatogram. •

            •   If the chromatogram of any standard needs to be replotted
                electronically to meet these  requirements, both the  initial
                chromatogram and the replotted chromatogram must  be  submitted
                in the data package.

9.2.6    Corrective Action for Initial Calibration

9.2.6.1     If the technical acceptance criteria for the initial calibration
            are not met,  inspect the system for problems.   It may be necessary
            to change the column,  bake out the detector,  clean the injection
            port,  or take other corrective actions to achieve the acceptance
            criteria.

9.2.6.2     Contamination should be suspected as a cause if the detector
            cannot achieve acceptable linearity using this method.  In the
            case of low level contamination,  baking out the detector at an
            elevated temperature (350 °C)  should be sufficient to achieve
            acceptable performance.   In the case of heavy contamination,
            passing hydrogen through the detector for 1-2 hours at an elevated
            temperature may correct the problem.  In the case of severe
            contamination, the detector may require servicing by the BCD
            manufacturer.  DO NOT OPEN THE DETECTOR.  THE BCD CONTAINS
            RADIOCHEMICAL SOURCES.

9.2.6.3     If a laboratory cleans out a detector using an elevated
            .temperature,  the BCD electronics must be turned off during the
            bake out procedure.

9.2.6.4     After bake out or hydrogen reduction, the detector must be
            recalibrated using the initial calibration sequence.

9.2.6.5     Initial calibration technical acceptance criteria must be met
            before any samples, including MS/MSD or required blanks, are
            analyzed.  Any samples or required blanks analyzed before the
            initial calibration technical acceptance criteria have been met
            will, require reanalysis at no additional cost to the Agency.
                                   D-27/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 9
Calibration and Standardization
Calibration Verification
9.3   Calibration Verification

9.3.1    Summary of Calibration Verification

         Three types of analyses are used to verify the calibration and
         evaluate instrument performance.  The analyses of instrument blanks,
         PEMs, and the midpoint concentration of Individual Standard Mixtures
         A and B constitute the continuing calibration.  Sample data are not
         acceptable unless bracketed by acceptable analyses of instrument
         blanks, PEMs, and both Individual Standard Mixtures A and B.

9.3.2    Frequency of Calibration Verification

9.3.2.1     An instrument blank and the PEM must bracket one end of a 12-hour
            period during which sample data are collected, and a second
            instrument blank and the midpoint concentration of Individual
            Standard Mixtures A and B must bracket the other end of the 12-
            hour period.

9.3.2.2     For the 12-hour period immediately following the initial
            calibration sequence, the instrument blank and the PEM that are
            the last two steps in the initial calibration sequence bracket the
            front end of that 12-hour period.  The injection of the instrument
            blank starts the beginning of that 12-hour period  (see Section
            10.2.2.1).  Samples may be injected for 12 hours from the
            injection of the instrument blank.  The three injections
            immediately after that 12-hour, period must be an instrument blank,
            Individual Standard Mixture A, and Individual Standard Mixture B.
            The instrument blank must be analyzed first, before either
            standard.  The Individual Standard Mixtures may be analyzed in
            either order  (A,B or B,A).

9.3.2.3     The analyses of the instrument blank and Individual Standard
            Mixtures A and B immediately following one 12-hour period may be
            used to begin the subsequent 12-hour period, provided that they
            meet the acceptance criteria in Section 9.3.5.  In that instance,
            the subsequent 12-hour period must be bracketed by the acceptable
            analyses of an instrument blank and a PEM, in that order.  Those
            two analyses may in turn be used to bracket the front end of yet
            another 12-hour period.  This progression may continue every 12
            hours until such time as any of the instrument blanks, PEMs, or
            Individual Standard Mixtures fails to meet the acceptance criteria
            in Sections 9.3.5.  The 12-hour time period begins with the
            injection of the instrument blank.

9.3.2.3.1      Standards  (PEM or Individual Standard Mixtures), samples and
               required blanks may be injected for 12 hours from the time of
               injection of the instrument blank.

9.3.2.4     If more than 12 hours have elapsed since the injection of the
            instrument blank that bracketed a previous 12-hour period, an
            acceptable instrument blank and PEM must be analyzed in order to
            start a new sequence.  This requirement applies even if no

                                   D-28/PEST                            OLM04.2

-------
                                    Exhibit D Pesticides/Aroclors -- Section 9
                                               Calibration and Standardization
                                                      Calibration Verification
            analyses were performed since that standard was injected.

9.3.2.5     The requirements for running the instrument blanks,  PEM, and
            Individual Standard Mixtures A and B are waived when no samples,
            dilutions, reanalyses, method/sulfur blanks, MS/MSD, and
            multicomponent analytes for the 72-hour confirmation requirement
            are analyzed during that 12-hour period.  To resume analysis,
            using the existing initial calibration, the Contractor first must
            analyze an instrument blank and PEM which meet the technical
            acceptance criteria.

9.3.2.6     If the entire 12-hour period is not required for the analyses of
            all samples to be reported, the sequence must be ended with either
            the instrument blank/PEM combination or the instrument
            blank/Individual Standard Mixtures A and B combination, whichever
            was due to be performed at the end of the 12-hour period.

9.3.3    Procedure for Calibration'Verification

9.3.3.1     Analyze the PEM, instrument blank, and the midpoint concentration
            of Individual Standard Mixtures A and B at the required
            frequencies  (Section 9.3.2).

9.3.3.2     All standards and blanks must be at ambient temperature at the
            time of preparation and analysis.

9.3.4    Calculations for Calibration Verification

9.3.4.1     For each  analysis of the PEM used to demonstrate continuing
            calibration, calculate the percent difference between the amount
            of each analyte  (including the surrogates)  found in the PEM and
            the nominal amount using Equations 5 and 9.

9.3.4.2     Calculate the percent breakdown of DDT and endrin, and  the
            combined  breakdown  in each PEM analyzed using Equations 5, 6, 7,
            and 8.

9.3.4.3     For each  analysis of the midpoint concentration of  Individual
            Standard  Mixtures A and B used to demonstrate continuing
            calibration, calculate the percent difference between the amount
            of. each analyte  (including  the surrogates)  found in the standard
            mixture and  the  nominal amount using Equations 5 and 9.  Do not
            attempt to calculate  the breakdown of  Endrin and DDT in the
            Individual Standard Mixtures, as these  standards contain the
            breakdown products  as well  as the parent compounds.

9.3.5    Technical Acceptance Criteria  for Calibration Verification

         All  calibration verification  technical acceptance criteria apply
         independently to each  column.  Each column must meet criteria.

9.3.5.1     The PEMs, Individual  Standard Mixtures A and B and  instrument
            blanks must  be  analyzed at  the required frequency on a  GC/EC

                                   D-29/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 9
Calibration and Standardization
Calibration Verification
            system that has met the initial calibration technical acceptance
            criteria.

9.3.5.2     All single component pesticides and surrogates in the PEMs used to
            demonstrate continuing calibration must be greater than or equal
            to 90.0 percent resolved.  The resolution between any two adjacent
            peaks in the midpoint concentrations of Individual Standard
            Mixtures A and B used to demonstrate initial calibration must be
            greater than or equal to 90.0 percent.

9.3.5.3     The absolute retention time .for each of the single component
            pesticides and surrogates in the PEMs and midpoint concentration
            of the Individual Standard Mixtures used to demonstrate continuing
            calibration must be within the retention time window determined
            from the three-point initial calibration in Section 9.2.4.3.

9.3.5.4     The percent difference between the calculated amount and the
            nominal amount (amount added) for each of the single component
            pesticides and surrogates in the PEM must be greater than or equal
            to -25.0 percent and less than or equal to 25.0 percent.

9.3.5.5     The percent difference between the calculated amount and the
            nominal amount (amount added) for each of the single component
            pesticides and surrogates in the INDA and INDB that have been used
            as calibration verification must be greater than or equal to -25.0
            percent and less than or equal to 25.0 percent.

9.3.5.6     The percent breakdown of DDT in the PEM must be less than or equal
            to 20.0 percent on each column.  The percent breakdown of endrin
            in the PEM must be less than or equal to 20.0 percent on each
            column.  The combined breakdown of both DDT and endrin must be
            less than or equal to 30.0 percent on each column.

9.3.5.7     All instrument blanks must meet the technical acceptance criteria
            in Section 12.1.4.4.

9.3.5.8     The identification of single component pesticides by gas
            chromatographic methods is based primarily on retention time data.
            Since the retention time of the apex of a peak can only be
            verified from an on-scale chromatogram, the following requirements
            must be met for continuing calibration to be acceptable.

9 . 3 . 5 . 8.. 1      The chromatograms that result from the analyses of the PEM and
               the Individual Standard Mixtures must display the single
               component analytes present in each standard at greater than 10
               percent of full scale but less than 100 percent of full scale.

9.3.5.8.2      For any PEM, Individual Standard Mixture or blank, the baseline
               of the chromatogram must return to below 50 percent of full
               scale before the elution time of alpha-BHC, and return to below
               25 percent of full scale after the elution time of alpha-BHC
               and before the elution time of decachlorobiphenyl.


                                   D-30/PEST                            OLM04.2

-------
                                    Exhibit D Pesticides/Aroclors -- Section 9
                                               Calibration and Standardization
                                                      Calibration Verification
9.3.5.8.3      If a chromatogram is replotted electronically to meet these
               requirements, the scaling factor used must be displayed on the
               chromatogram.

9.3.5.8.4      If the chromatogram of any standard or blank needs to be
               replotted electronically to meet these requirements,  both the
               initial chromatogram and the replotted chromatogram(s)  must be
               submitted in the data package.

9.3.6    Corrective Action  for Calibration Verification

9.3.6.1     If the technical acceptance criteria for the calibration
            verification are not met, inspect the system for problems and take
            corrective action to achieve the acceptance criteria.

9.3.6.2     Major corrective actions such as replacing the GC column or baking
            out the detector will require that a new initial calibration be
            performed that  meets the technical acceptance criteria in Section
            9.2.5.  _

9.3.6.3     Minor corrective actions may not require performing a new initial
            calibration, provided that a new analysis of the standard  (PEM or
            Individual Standard Mixture) that originally failed the criteria
            and an associated instrument blank immediately after the
            corrective action do meet all the acceptance criteria.

9.3.6.4     If a PEM or Individual Standard Mixture does not meet technical
            acceptance criteria listed above, it must be reinjected
            immediately.   If the second injection of the PEM or Individual
            Standard Mixture meets the criteria, sample analysis may continue.
            If the second  injection does not meet the criteria, all data
            collection must be stopped.  Appropriate corrective action must be
            taken, and a new initial calibration sequence must be run before
            more sample data are collected.

9.3.6.5     If an instrument blank does not meet the technical acceptance
            criteria listed in Section 12.1.4.4, all data collection must be
            stopped.  Appropriate corrective action must be taken to clean out
            the  system, and an acceptable instrument blank must be analyzed
            before more sample data are collected.

9.3.6.6     Analysts are reminded that running an instrument blank and a PEM
            or Individual  Standard Mixtures once every  12 hours is the minimum
            contract requirement.  Late eluting peaks may carry over from one
            injection to the next if highly complex samples are analyzed or if
            the  GC conditions are unstable.  Such carryover is unacceptable.
            Therefore,  it  may be necessary to run instrument blanks and
            standards more often to avoid discarding data.

9.3.6.7     If a successful instrument blank and PEM cannot be run after an
            interruption in analysis  (Section 9.3.2.5),  an acceptable  initial
            calibration must be run before sample data  may be collected.  All
            acceptable  sample analyses must be preceded and followed by

  -. . •                  .            D-31/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 9
Calibration and Standardization
Calibration Verification
            acceptable standards and instrument blanks, as described in
            Section 9.3.2.
9.3.6.8     Calibration verification technical acceptance criteria must be met
            before any samples, including MS/MSD and required blanks are
            reported.  Any samples, including MS/MSD and required blanks
            associated with a calibration verification which did not meet the
            technical acceptance criteria will require reanalysis at no
            additional cost to the Agency.
                                   D-32/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                            Sample Preparation

10.0  PROCEDURE

10.1  Sample Preparation

10.1.1   If insufficient sample amount (less than 90% of the required amount)
         is received to perform the analyses, the Contractor shall contact SMO
         to apprise them of the problem.   SMO will contact the Region for
         instructions.   The Region will either require that no sample analyses
         be performed or will require that a reduced volume be used for the
         sample analysis.  No other changes in the analyses will be permitted.
         The Contractor shall document the Region's decision in the SDG        '
         Narrative.

10.1.2   If multiphase samples  (e.g., a two-phase liquid sample, oily
         sludge/sandy soil sample) are received by the Contractor, the
         Contractor shall contact SMO to apprise them of the type of sample
         received.  SMO will contact the Region.  If all phases of the sample
         are amenable to analysis, the Region may require the Contractor to do
         any of the following:

         •   Mix the sample and analyze an aliquot from the homogenized
             sample.

         •   Separate the phases of the sample and analyze each phase
             separately.  SMO will provide EPA sample numbers for the
             additional phases.

         •   Separate the phases and analyze one or more of the phases, but
             not all of the phases.  SMO will provide EPA sample numbers for
             the additional phases, if required.

         •   Do not analyze the sample.

10.1.2.1    If all  of the phases are not amenable to analysis  (i.e., outside
            scope), the Region may require the Contractor to do any of the
            following:

            •    Separate the phase(s)  and analyze the phase(s)  that is
                 amenable to analysis.   SMO will provide EPA sample numbers
                 for the additional phases,  if required.

            •    Do not analyze the sample.

10.1.2.2    No other change in  the analyses  will be permitted.  The Contractor
            shall  document the  Region's decision in the SDG Narrative.

10.1.3   Extraction of Water  Samples

         Water samples may be extracted by either a separatory  funnel
         procedure or a  continuous liquid-liquid extraction procedure.   If an
         emulsion  prevents acceptable  solvent recovery with the separatory
         funnel procedure, continuous  liquid-liquid extraction must be
         employed.


                                   D-33/PEST                           OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure
Sample Preparation

10.1.3.1    Separatory Funnel Extraction

10.1.3.1.1     Measure out each 1 L sample aliquot in a separate graduated
               cylinder.  Measure and record pH of the sample with wide range
               pH paper and adjust the pH to between 5 and 9 with 10 N sodium
               hydroxide or concentrated sulfuric acid, if required.  Samples
               requiring pH adjustment must be noted in the SDG Narrative.
               Place the sample into a 2 L.separatory funnel.

10.1.3.1.2     Using a syringe or a volumetric pipet, add 1.0 mL of the
               surrogate solution to all water samples.

10.1.3.1.3     Rinse the graduated cylinder with 30 mL of methylene chloride
               and transfer the rinsate to the separatory funnel.  If the
               sample was received in a 1 L container, rinse the empty
               container with 30 mL of methylene chloride and add rinsate to
               the separatory funnel.  If the sample container is not rinsed,
               then add another 30 mL of methylene chloride to the separatory
               funnel and extract the sample by shaking the funnel for two
               minutes, with periodic venting to release excess pressure.
               NOTE:  The total volume of solvent used for extraction is 60
               mL.  Allow the organic layer to separate from the water phase
               for a minimum of 10 minutes.  If the emulsion interface between
               layers is more than one-third the volume of the solvent layer,
               the analyst must employ mechanical techniques to complete the
               phase separation.  The optimum technique depends upon the
               sample, and may include stirring, filtration of the emulsion
               through glass wool, centrifugation or other physical means.
               Drain the methylene chloride into a 250 mL Erlenmeyer flask.

10.1.3.1.4     Add a second 60 mL volume of methylene chloride to the
               separatory funnel and repeat the extraction procedure a second
               time, combining the extracts in the Erlenmeyer flask.  Perform
               a third extraction in the same manner.

10.1.3.2    Continuous Liquid-Liquid Extraction

10.1.3.2.1     Continuous Liquid-Liquid Extraction Without Hydrophobic
               Membrane

10.1.3.2.1.1      Follow manufacturer's instructions for set-up.

10.1.3.2.1.2     . Add methylene chloride to the bottom of the extractor and
                  fill it to a depth of at least one inch above the bottom
                  sidearm.

10.1.3.2.1.3      Measure out each 1 L sample aliquot in a separate, clean
                  graduated cylinder; transfer the aliquot to the continuous
                  extractor.  Measure the pH of the sample with wide range pH
                  paper and record pH.  Adjust the pH to between 5 and 9 with
                  10 N sodium hydroxide or concentrated -sulfuric acid, as
                  required.  Samples requiring the pH adjustment must be noted
                  in the SDG Narrative.  NOTE:  With some samples, it may be
                  necessary to place a layer of glass wool between the

                                   D-34/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                            Sample Preparation

                  methylene chloride and the water layer in the extractor to
                  prevent precipitation of suspended solids into the methylene
                  chloride during extraction.

10.1.3.2.1.4      Using a syringe or volumetric pipet,  add 1.0 mL of the
                  surrogate standard spiking solution (Section 7.2.4.1) into
                  the sample and mix well.

10.1.3.2.1.5      Rinse the graduated cylinder with 50 mL of methylene
                  chloride and transfer the rancid to the continuous
                  extractor.  If the sample was received in a 1 L container,
                  rinse the empty container with 50 mL of methylene chloride
                  and add the rancid to the continuous extractor.

10.1.3.2.1.6      Add sufficient methylene chloride to the continuous
                  extractor to ensure proper solvent cycling during operation.
                  Adjust the drip rate to 5 to 15 mL/minute (recommended);
                  optimize the extraction drip rate.  Extract for a minimum of
                  18 hours.  NOTE:  When.a minimum drip rate of 10-15
                  mLs/minute is maintained throughout the extraction, the
                  extraction time may be reduced to a minimum of 12 hours.
                  Allow to cool, then detach the distillation flask.  Proceed
                  to Section 10.1.4.

10.1.3.2.1.7      NOTE:  Some continuous liquid-liquid extractors are also
                  capable of concentrating the extract within the extraction
                  set-up.  Follow the manufacturer's instructions for
                  concentration when using this type of extractor.

10.1.3.2.2     Continuous Liquid-Liquid Extraction With Hydrophobic Membrane

10.1.3.2.2.1      Follow the manufacturer's  instructions for set-up.

10.1.3.2.2.2      Measure out each  1 L sample aliquot in a separate, clean
                  graduated cylinder.  If the sample was received in a  1 L
                  container, rinse  the empty container with 50 mL of methylene
                  chloride and add  the rinsate to the continuous extractor.
                  If the sample was not received in a 1 L container, add 50 mL
                  of methylene chloride to the continuous extractor.   Slowly
                  transfer the aliquot to the continuous extractor.  Measure
                  the pH of the sample with  wide range pH paper and record pH.
                  Adjust the pH to  between 5 and 9 with 10 N sodium hydroxide
                  or concentrated sulfuric acid, as required.  Samples
                  requiring the pH  adjustment must be noted in the SDG
                  Narrative.

10 .1.3.. 2 .2 . 3      Using a  syringe or volumetric pipet, add 1.0 mL of the
                  surrogate standard spiking solution  (Section 7.2.4.1)  into
                  the  sample and mix well.

10.1.3.2.2.4      Rinse the graduated cylinder with 50 mL of methylene
                  chloride  and  transfer the  rinsate to the continuous
                  extractor.
                                   D-35/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure
Sample Preparation

10.1.3.2.2.5      Add sufficient methylene chloride to the continuous
                  extractor to ensure proper solvent cycling during operation.
                  Adjust the drip rate to 15 mL/minute (recommended); optimize
                  the extraction drip rate.  Extract for a minimum of 6 hours.
                 . (NOTE:  Due to the smaller volume of solvent used during the
                  extraction process, some sample matrices (e.g., oily
                  samples, samples containing a high concentration of
                  surfactants) may create an emulsion which will consume the
                  solvent volume, preventing the efficient extraction of the
                  sample.  When this occurs, add additional solvent to assure
                  efficient extraction of the sample, and extend the
                  extraction time for a minimum of 6 hours.  If the sample
                  matrix prevents the free flow of solvent through the
                  membrane, then the non-hydrophobic membrane continuous
                  liquid-liquid type extractor must be used.)  Allow to cool,
                  then detach the distillation flask.

10.1.3.2.2.6      NOTE:  Some continuous liquid-liquid extractors are also
                  capable of concentrating the extract within the extraction
                  act-up.  Follow the manufacturer's instructions for
                  concentration when using this type of extractor.  Using the
                  hydrophobia membrane, it may not be necessary to dry the
                  extract with sodium sulfate.

10.1.3.2.3     NOTE:  If low surrogate recoveries occur, assure 1) the
               apparatus was properly assembled to prevent leaks; 2) the drip
               rate/solvent cycling was optimized; and 3) there was proper
               cooling for condensation of solvent.

10.1.3.2.4     NOTE:  Alternate continuous liquid-liquid extractor types that
               meet the requirements of the SOW may also be used.  If using
               alternate extractors or design types, follow the manufacturer's
               instructions for set-up.  Optimize the extraction procedure.

10.1.4   Extract Drying

10.1.4.1    Assemble a Kuderna-Danish  (K-D) concentrator by attaching a 10 mL
            concentrator tube to a 500 mL evaporative flask.  Other
            concentration devices or techniques may be used in place of the
            K-D if equivalency is demonstrated for all the pesticide/Aroclor
            target compounds listed 'in Exhibit C.

10.1.4.2    Pour the extract through a drying column containing about 10 cm of
            anhydrous granular sodium sulfate and collect the extract in the
            K-D concentrator.  Rinse the Erlenmeyer flask and the column with
            at least two additional 20 to 30 mL portions of methylene chloride
           • to complete the quantitative transfer.

10.1.5   Soil/Sediment Samples

         Mix samples thoroughly, especially composited samples.  Discard any
         foreign objects such as sticks, leaves, and rocks.  Also, decant and
         discard any standing aqueous phase.


                                   D-36/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                            Sample Preparation
10.1.5.1    pH Determination
10.1.5.1.1     Transfer 50 g of soil/sediment to a 100 mL beaker.   Add 50 mL
               of water and stir the solution with a magnetic stirrer for 1
               hour.  Determine the pH of the sample by using a combination
               glass electrode and pH meter while the sample is stirred.
               Report the pH value on the appropriate data sheet.   If the pH
               of the soil/sediment is > 9 or < 5, document any subsequent
               problems In the analysis related to pH in the SDG Narrative,
               but do not attempt to adjust the pH of the sample.   Discard the
               portion of the sample used for pH determination.

10.1.5.1.2     NOTE:  If insufficient volume of soil/sediment is received, use
               a smaller 1:1 ratio of grams of sample to mLs of water for the
               pH determination, and note in the SDG Narrative.

10.1.5.2    Percent Moisture

            Weigh 5 to 10 g of the soil/sediment to the nearest 0.01 g into a
            tared crucible or aluminum weighing pan.  Determine the weight
            percent volatilized by drying overnight at 105 °C  (hereafter
            referred to as percent moisture).  After the sample is dry, remove
            the sample and pan and allow them to cool in a desiccator before
            weighing.  Calculate the percent moisture according to Equation 11
            below.  Concentrations of individual analytes will be reported
            relative to the dry weight of soil/sediment.  CAUTION:  Gases
            volatized from some soil/sediment samples may require that this
            drying procedure be carried out  in a hood.
            EQ. 11

                       grams of wi
                                  grams of wet sample
......      grams of wet sample - grams of dry sample   , „„
%Moisture = —	—	— x 100
 10.1.5.3    Soil/Sediment  Extraction

            The procedure  described below is  for the  extraction of
            soil/sediment  samples by  sonication.   The Contractor may also use
            Automated Soxhlet Extraction  (SW-846 Method 3541 Revision  0,
            September 1994)  or Pressurized Fluid Extraction  (Sff-846  Method
            3545  Revision  0,  December 1996) techniques for soil/sediment
            samples.   The  above SW-846 methods  are provided as reference only
            and  the laboratory supplied Standard Operating Procedures  must be
            accepted by the Agency before the laboratory can utilize these
            methods (See Exhibit E for required IPR studies).  The
            requirements of this SOW  must be  met at all times  (i.e., original
            sample weight).   As applicable, follow manufacturer's instructions
            for  use of all extraction equipment.   If  one of the above
            alternative extraction procedures is used,  the Contractor  must
            maintain documentation of the procedure utilized and document its
            equivalence to the sonication procedure described below.   If the
            sample weight  must be adjusted to utilize one of the alternative
            extraction procedures, the Contractor shall immediately  contact

                                   D-37/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure
Sample Preparation

            SMO to inform them of the problem.  SMO will contact the Region
            for instructions.

            Note: All soil/sediment samples in a Case must be extracted by the
            same procedure.

io.1.5.3.1     Tune the sonicator according to the manufacturer's directions
               prior to extracting samples by this procedure.

10.1.5.3.2     Weigh approximately 30 g of sample  (to the nearest 0.1 g) into
               a 250 or 400 mL beaker and add 60 g of anhydrous sodium sulfate
               (granular).

10.1.5.3.3     Add 2.0 mL of surrogate solution to all soil/sediment samples
               by using a volumetric pipet or a syringe.  Mix the sample well.
               The sample and the added sodium sulfate should be a
               homogeneous, granular mixture at this point.  Twice as much of
               the surrogate solution is added to soil/sediment samples than
               to water samples.

10.1.5.3.4     Immediately add 80 to 100 mL of 1:1 methylene chloride/acetone
               to the sample.

10.1.5.3.5     Place the bottom surface of the sonicator probe about M inch
               below the surface of the solvent but above the sediment layer.

10.1.5.3.6     Sonicate for 3 minutes using a 3/4 inch disrupter horn at full
               power  (output control knob at 10) with pulse on and percent
               duty cycle knob set at 50 percent.  Do not use a microtip.
               NOTE:  These settings refer to the Model W-385.  When using a
               sonicator other then Model W-385, refer to the instructions
               provided by the manufacturer for appropriate output settings.

10.1.5.3.7     The extracted sample can be filtered by using gravity or vacuum
               filtration.

10.1.5.3.8     For gravity filtration prepare a filtration/drying bed by
               placing a plug of glass wool in the neck of a 10 cm powder
               funnel and filling the funnel to approximately half its depth
               (4 or 5 cm) with anhydrous sodium sulfate  (80-100 g).  Decant
               the extract through the packed funnel and collect it in a
               500 mL evaporative  (K-D) flask attached to a concentrator tube.

10.1.5.3.9     For vacuum filtration, use Whatman No. 41 paper in the Buchner
               funnel.  Pre-wet the paper with methylene chloride/acetone  .
               before decanting the solvent.

10.1.5.3.10    Repeat the extraction two more times with additional 80 to
               100 mL portions of the 1:1 methylene chloride/acetone.  Before
               each extraction, thoroughly mix the solid residue and make
               certain that the sodium sulfate is free flowing and not a
               consolidated mass.  As required, break up large lumps with a
               clean spatula.  Decant and filter the extraction-solvent after
               each sonication by using the same funnel described above.

                                   D-38/PEST                    '        OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                     Concentrating the Extract

               After the final sonication, pour the entire sample into the,
               funnel and rinse the beaker and funnel with 60 mL of 1:1
               methylene chloride/acetone.

10.1.6   Concentrating the Extract

10.1.6.1    Concentration by K-D

            Other concentration devices or techniques may be used in place of.
            the K-D if equivalency is demonstrated for all the
            .pesticide/Aroclor target compounds listed in Exhibit C.

10.1.6.1.1     Add one or two clean boiling chips to the evaporative flask and
               attach a three-ball Snyder column.  Pre-wet the Snyder column
               by adding about 1 mL of methylene chloride to the top of the
               column.  Place the K-D apparatus on .a hot water bath  (60 °C to
               70 °C recommended) so that the concentrator tube is partially
               immersed in the hot water  and the entire lower rounded surface
               of the flask is bathed with hot vapor.  Adjust the vertical
               position of the apparatus  and the water temperature as required
               to complete the concentration in 15 to 30 minutes.  At the
               proper rate of distillation, the balls of the column will
               actively.chatter, but the  chambers will not flood with
               condensed solvent.  When the apparent volume of liquid reaches
               3, to 5 mL for water samples  (and less than 10 mL for
               soil/sediment samples), remove the K-D apparatus.  Allow it to
               drain and cool for at least 10 minutes.  DO NOT ALLOW THE
               EVAPORATOR TO GO DRY.

10.1.6.1.2     For water extracts which do not require GPC cleanup, and for
               water and soil/sediment extracts which have been through the
               GPC cleanup step, proceed  with the hexane exchange described in
               Section 10.1.6.2.

10.1.6.1.3     For water extracts which require GPC cleanup, remove the Snyder
               column, rinse the flask and its lower joint, collect the
               rinsate in the concentrator tube, and adjust the volume to
               10 mL with methylene chloride.  Proceed to Section 10.1.8.1.

10.1.6.1.4     For soil/sediment extracts that have not been cleaned-up using
               GPC, it is absolutely necessary to further reduce the volume of
               all soil/sediment extracts to 1 mL in order to remove most of
               the acetone.  This is best accomplished using the nitrogen
               evaporation technique  (Section 10.1.7.2).  The presence of
               acetone will cause a dead  volume to develop in the GPC column
               and thus will cause loss of surrogates and analytes during GPC
               cleanups.  Adjust the soil/sediment extract volume to 10 mL
               with methylene chloride.   Proceed to Section 10.1.8.1 for
               mandatory GPC.    . .

10.1.6.2     Solvent Exchange into Hexane

             This procedure applies to both extracts of water samples and
             extracts  of soil/sediment samples.

                                   Dr39/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure
Concentrating the Extract
10.1.6.2.1     With the extract in a K-D apparatus, remove the Snyder column,
               add 50 mL of hexane and a new boiling chip, and reattach the
               Snyder column.  Pre-wet the column by adding about 1 mL of
               hexane to the top.  Concentrate the solvent extract as
               described previously  (Section 10.1.6.1), but increase the
               temperature of the water bath (to between 80 and 90 °C
               recommended).  When the apparent volume of liquid reaches 3 to
               5 mL, remove the K-D apparatus and allow it to drain and cool
               for at least 10 minutes.  DO NOT ALLOW THE EVAPORATOR TO GO
               DRY.

10.1.6.2.2     Remove the Snyder column; using 1 to 2 mL of hexane, rinse the
               flask and its lower joint into the concentrator tube.  Complete
               quantitative transfer of the extract to a 10 mL vial by using
               hexane.

10.1.6.2.3     For samples which have not been subjected to GPC cleanup,
               adjust the volume of the hexane extract to 10 mL.  For samples
               which have been subjected to GPC cleanup, concentrate the
               hexane extract tb 5 mL using a micro Snyder column or nitrogen
               evaporation, as described in Section 10.1.7.1 or 10.1.7.2.
               Proceed to Section 10.1.8.2 for Florisil cartridge cleanup.

10.1.7   Final Concentration of Extract

         Two different techniques are permitted to concentrate the extract to
         volume before Florisil cleanup or extract volume before instrumental
         analysis.  They are the. Micro Snyder Column and Nitrogen Evaporation
         Technique.

10.1.7.1    Micro Snyder Column Concentration

            Add another one or two clean boiling chips to the concentrator
            tube and attach a two-ball micro Snyder column.  Pre-wet the
            Snyder column by adding about 0.5 mL of hexane to the top of the
            column.  Place the K-D apparatus in a hot water bath (80 °C to 90
            °C recommended) so that the concentrator tube is partially
            immersed in the hot water.  Adjust the vertical position of the
            apparatus and the water temperature as required to complete the
            concentration in 5 to 10 minutes.  At the proper rate of
            distillation, the balls of the column will actively chatter, but
            the chambers will not flood with condensed solvent.  When the
            apparent volume of liquid reaches 0.5 mL, remove the K-D apparatus
            from the water bath and allow it to drain for at least 10 minutes
            while cooling.  Remove the Snyder column and rinse its flask and
            lower joint into the concentrator tube with 0.2 mL of hexane.
            Adjust the final volume with hexane to 1 or 2 mL (see Sample
            Cleanup by Florisil Cartridge, Section 10.1.8.2.3).
                                   D-40/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                            Cleanup Procedures

10.1.7.2    Nitrogen Evaporation Technique (taken from ASTM Method D 3086)

10.1.7.2.1     Place the Concentrator tube in a warm water bath (30 °C to
               35 °C recommended)  and evaporate the solvent volume to the
               final volume by blowing a gentle stream of clean, dry nitrogen
               (filtered through a column of activated carbon) onto the
               solvent.  DO NOT ALLOW THE EXTRACT TO GO TO DRYNESS.  Adjust
               the soil/sediment extract volume to 10 mL with methylene
               chloride.  Proceed to Section 10.1.8 for mandatory GPC cleanup
               procedures.

10.1.7.2.2     Gas lines from the gas source to the evaporation apparatus must
               be stainless steel, copper, or Teflon tubing.  Plastic tubing
               must not be used between the carbon trap and the sample as it
               may introduce interferences.  The internal wall of new tubing
               must be rinsed several times with hexane and then dried prior
               to use.

10.1.7.2.3     During evaporation, the tube solvent level must be kept below
  .  •           the water level of the bath.

10.1.8   Cleanup Procedures

         There are three cleanup procedures specified in this method:  GPC,
         Florisil cartridge, and sulfur cleanup.  GPC must- be performed  for
         all soil/sediment extracts.  GPC must be performed for water extracts
         that contain higher molecular weight contaminants that interfere with
         the analysis of the target  analytes.  Florisil cartridge  cleanup is
         mandatory for all extracts.  Sulfur cleanup must  be performed for all
         sample extracts contaminated with sulfur.  Blanks and matrix spike
         and matrix spike duplicate  samples must be subjected to the same
         cleanup as the unspiked samples.

10.1.8.1    Sample Cleanup by Gel Permeation Chromatography  (GPC)

10.1.8.1.1     Introduction

               Gel Permeation Chromatography  (GPC)  is a size  exclusion cleanup
               procedure using organic  solvents  and hydrophobic gels in  the
               separation of natural (and synthetic) macromolecules.  The
               packing  gel  is porous and  is  characterized  by  the  range or
               uniformity  (exclusion range)  of  that pore size.  In the choice
               of gels,  the exclusion range must be larger than the molecular
               size of  the molecules to be  separated.

10.1.8.1.2     GPC Column  Preparation

               The  instructions  listed  below for GPC column preparation  are
               for Bio  Beads.  Alternate  column packings may  be used  if  1) the
               column packings have  equivalent  or  better performance  than the
               Bio Beads  and meet  the technical  acceptance criteria  for  GPC
               calibration  and GPC  calibration checks,  and 2).the  column
               packings do  not  introduce  contaminants/artifacts into  the
               sample which interfere with the  analysis of the  pesticide

                                   D-41/PEST                             OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure
Cleanup Procedures

               compounds.  Follow the manufacturer's instructions for
               preparation of the GPC column packing.

10.1.8.1.2.1      Weigh out 70 g of Bio Beads  (SX-3).  Transfer them to a 1 L
                  bottle with a Teflon-lined cap or a 500 mL separatory funnel
                  with a large bore stopcock, and add approximately 300 mL of
                  methylene chloride.  Swirl the container to ensure the
                  wetting of all beads.  Allow the beads to swell for a
                  minimum of 2 hours.  Maintain enough solvent to cover the
                  beads sufficiently at all times.  If a guard column is to be
                  used, repeat the above with 5 g of Bio Beads in a 125 mL
                  bottle or a beaker, using 25 mL of methylene chloride.

10.1.8.1.2.2      Turn the column upside down from its normal position and
                  remove the inlet bed support plunger  (the inlet plunger is
                  longer than the outlet plunger).  Position and tighten the
                  outlet bed support plunger as near the end as possible, but
                  no closer than 5 cm  (measured from the gel packing to the
                 . collar).        ;

10.1.8.1.2.3      Raise the end of the outlet tube to keep the solvent in the
                  GPC column, or close the column outlet stopcock.  Place a
                  small amount of solvent in the column to minimize the
                  formation of air bubbles at the base of poured column
                  packing.

10.1.8.1.2.4      Swirl the bead/solvent slurry to get a homogeneous mixture
                  and, if the wetting was done in a quart bottle, quickly
                  transfer it to a 500 mL separatory funnel with a large bore
                  stopcock.  Drain the excess methylene chloride directly into
                  the waste beaker, and then start draining the slurry into
                  the column by placing the separatory funnel tip against the
                 .column wall.  This will help to minimize bubble formation.
                  Swirl occasionally to keep the slurry homogeneous.  Drain
                  enough to fill the column.  Place the tubing from the column
                  outlet into a waste beaker below the column, open the
                  stopcock (if attached), and allow the excess solvent to
                  drain.  Raise the tube to stop the flow, and close the
                  stopcock when the top of the gel begins to look dry.  Add
                  additional methylene chloride to just rewet the gel.

10.1.8.1.2.5      Wipe any remaining beads and solvent from the inner walls of
                  the top of the column with a laboratory tissue.  Loosen the
                  seal slightly on the other plunger assembly  (long plunger)
                  and insert it into the column.  Make the seal just tight
                  enough so that any beads on the glass surface will be pushed
                  forward, but loose enough so that the plunger can be pushed
                  forward.

                  CAUTION:  Do not tighten the seal if beads are between the
                  seal and the- glass surface because this can damage the seal
                  and cause leakage.
                                   D-42/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                            Cleanup Procedures

10.1.8.1.2.6      Compress the column as much as possible without applying
                  excessive force.  Loosen the seal and gradually pull out the
                  plunger.-  Rinse and wipe off the plunger.  Slurry any
                  remaining beads and transfer them into the column.  Repeat
                  the step in Section 10.1.8.1.2.5 and reinsert the plunger.
                  If the plunger cannot be inserted and pushed in without
                  allowing beads to escape around the seal, continue
                  compression of the beads without tightening the seal, and
                  loosen and remove the plunger as described.  Repeat this
                  procedure until the plunger is inserted successfully.

10.1.8.1.2.7      Push the plunger until it meets the gel, then compress the
                  column bed about four centimeters.

10.1.8.1.2.8      Pack the optional 5 cm column with approximately 5 g of pre
                  swelled beads  (different guard columns may require different
                  amounts).  Connect the guard column to the inlet of the
                  analytical column.

10 .1. 8.. 1. 2 . 9      Connect the column inlet to the solvent reservoir  (reservoir
                  should be placed higher than the top of the column) and
                  place the column outlet tube in a waste container.  Placing
                  a restrictor in the outlet tube will force air out of the
                  column more quickly.  A restrictor can be made from a piece
                  of capillary stainless steel tubing of 1/16" OD x  0.010"  ID
                  x 2".  Pump methylene chloride through the column  at a rate
                  of 5 mL/min for one hour.

10.1.8.1.2.10     After washing  the column for at least one hour, connect the
                  column outlet  tube, without the restrictor, to the inlet
                  side of  the UV detector..  Connect the system outlet to the
                  .outlet side of the UV detector.  A restrictor  (same size  as
                  the one  in Section 10.1.8.1.2.9 above) in the outlet tube
                  from' the UV detector will prevent bubble formation which
                  causes a noisy UV baseline.  The restrictor will not affect
                  flow rate.  After pumping methylene chloride through the
                  column for an  additional 1-2 hours, adjust the inlet bed
                  support plunger until approximately 6-10 psi backpressure is
                  achieved.  Push the plunger in to increase pressure or
                  slowly pull outward to reduce pressure.

10.1.8.1.2.11     When the GPC column is not to be used for several  days,
                  connect  the column outlet line to the column inlet to
                  prevent  column drying and/or channeling.   If channeling
                  occurs,  the gel must be removed from the column,  reswelled,
                  and repoured as described above.  If drying occurs,
                  methylene chloride should be pumped through the  column until
                  the observed column pressure is constant and the  column
                  appears  wet.   Always recalibrate after  column drying has
                  occurred to verify retention volumes have not changed.
                  NOTE:  The description of solvent flow  rate and  column
                  pressure applies only to the ABC GPC apparatus.


                                   D-43/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure
Cleanup Procedures

                  Laboratories using equivalent equipment must develop the
                  parameters for their apparatus which give acceptable
                  performance as described in Section 10.1.8.1.3.

10.1.8.1.3     Calibration of GPC

10.1.8.1.3.1      Summary of GPC Calibration

10.1.8.1.3.1.1       The GPC calibration procedure is based on monitoring the
                     elution of standards with a UV detector connected to the
                     GPC column.

10.1.8.1.3.1.2       The UV detector calibration procedure described in
                     Section 10.1.8.1.3.3 is needed for the analyses of
                     organochlorine pesticides and Aroclors listed in Exhibit
                     C.  IT MUST NOT BE USED FOR THE ANALYSIS OF GC/MS
                     EXTRACTABLES OR OTHER ANALYTES WITHOUT A RECOVERY STUDY.

10.1.8.1.3.2      Frequency of GPC Calibration

                  Each GPC system must be calibrated upon award of the
                  contract, when the GPC calibration verification solution
                  fails to meet criteria, when the column is changed, when
                  channeling occurs, and once every seven days.

10.1.8.1.3.3      Procedure for GPC Calibration

                  Follow the manufacturer's instructions for operating the GPC
                  system.  Changes in pressure, solvent flow rate, and
                  temperature conditions can affect analyte retention times
                  and must be monitored.

10.1.8.1.3.3.1       Using a 10 mL syringe, load the calibration solution
                      (Section 7.2.4.3) onto the GPC.  Determine the elution
                     times for the phthalate, methoxychlor, and perylene.
                     Phthalate will elute first; perylene, last.

10.1.8.1.3.3.2      -Choose a "DUMP" time which removes > 85 percent of the
                     phthalate.  Choose a "COLLECT" time so that > 95 percent
                     of the methoxychlor is collected, and continue to collect
                     until just prior to the elution time of sulfur.  Use a
                     "WASH" time of 10 minutes.

                     NOTE:  The DUMP and COLLECT times must be adjusted to
                     compensate for the difference in volume of the lines
              ;       between the detector and the collection flask.

10.1.8.1.3.3.3       Reinject the calibration solution after appropriate
                     collect and dump cycles have been set, and the solvent
                     flow and column pressure have been established.

10.1.8.1.3.3.4       Measure and record the volume of collected GPC eluate in
                     a graduated cylinder.  The volume of GPC eluate collected
                     for each sample extract processed may be used to indicate

                                   D-44/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                            Cleanup Procedures

                     problems with the system during sample processing.

10.1.8.1.3.3.5       Analyze a GPC blank of methylene chloride.  Concentrate
                     the methylene chloride that passed through the system
                     during the collect cycle using Kuderna-Danish (K-D)
                     evaporator.  Exchange the solvent to hexane and analyze
                     the concentrate by GC/EC according to the procedure in
                     Section 10.2 (usual protocol).  Assuming that the blank
                     represents the extract from a 1 L water sample,  calculate
                     the analyte concentrations using Equation 13.

10.1.8.1.3.4      Technical Acceptance Criteria for GPC Calibration

10.1.8.1.3.4.1       The GPC system must be calibrated at the frequency
                     described in Section 10.1.8.1.3.2.  The UV trace must
                     meet the following requirements:

                     •  Peaks must be observed and should be symmetrical for
                        all compounds in the calibration solution.

                     •  Corn oil and phthalate peaks must exhibit  > 85 percent
                        resolution.

                     •  Phthalate and methoxychlor peaks must exhibit  > 85
                        percent resolution.

                     •  Methoxychlor and perylene peaks must exhibit >  85
                        percent resolution.

                     •  Perylene and sulfur peaks must not be saturated and
                        must exhibit > 90 percent baseline resolution.

10.1.8.1.3.4.2       The solvent flow rate and column pressure must be within
                     the manufacturer's specified ranges.

10.1.8.1.3.4.3       The retention times  for bis  (2-ethylhexyl) phthalate and
                     perylene must not vary more than ± 5 percent  between
                     calibrations.  If the retention time shift is > 5
                     percent, take corrective action.  Excessive retention
                     time shifts are caused by the following:

                     •  Poor laboratory temperature  control or  system  leaks.

                     •  An  unstabilized column that  requires pumping methylene
                        chloride through  it for several more hours or
                        overnight.

                     •  Excessive  laboratory  temperatures causing  outgassing
                        of  the  methylene  chloride.

10.1.8.1.3.4.4       The analyte concentrations in a GPC blank  must be less
                     than.the CRQL of any compound in Exhibit  C (Pesticides).

10.1.8.1.3.5      Corrective Action for GPC Calibration

                                   D-45/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure
Cleanup Procedures


10.1.8.1.3.5.1       If the flow rate and/or column pressure do not fall
                     within the manufacturer's specified ranges, a new column
                     should be prepared.

10.1.8.1.3.5.2       A UV trace that does not meet the criteria in Section
                     10.1.8.1.3.4 would also indicate that a new column should
                     be prepared.  It may be necessary to obtain a new lot of
                     Bio Beads if the column fails all the criteria.

10.1.8.1.3.5.3       If the GPC blank is equal to or exceeds the CRQL of any
                     compound in Exhibit C  (Pesticides), pump additional
                     methylene chloride through the system for 1-2'hours.
                     Analyze another GPC blank to ensure the system is
                     sufficiently clean.  Repeat the methylene chloride
                     pumping if necessary.

10.1.8.1.4  .   GPC Calibration Verification

10.1.8.1.4.1      Summary of GPC Calibration Verification

                  The GPC calibration must be routinely verified with two
                  check mixtures  (Section 7.2.4.3.3).

10.1.8.1.4.2      Frequency of GPC Calibration Verification

10.1.8.1.4.2.1       The calibration verification must be performed at least
                     once every 7 days  ("immediately following the GPC
                     Calibration) whenever samples, including matrix spikes,
                     matrix spike duplicates, and blanks are cleaned up using
                     the GPC.

10.1.8.1.4.2.2       Some samples may contaminate the SX-3 Bio Beads and
                     change the retention volume of the GPC column.
                     Therefore, system calibration and analyte recovery must
                     be checked whenever a sample causes significant
                     discoloration of the GPC column.  Even if no darkening is
                     visible, GPC calibration must be checked not less than-
                     once every seven days.

10.1.8.1.4.3      Procedure for GPC Calibration Verification

                  The instructions below are for a GPC injection loop of 5 mL.
                  If a 2 mL injection loop is used, the Contractor will adjust
                  the volume to 4 mL instead of 10 mL before injection extract
                  on the GPC.

10.1.8.1.4.3.1       The pesticide GPC calibration verification solution
                     contains the following six compounds in methylene
                     chloride: gamma-BHC  (Lindane), Heptachlor, and Aldrin
                     each at a concentration of 0.1 pg/mL for a 5 mL GPC loop
                      (0.25 ug/mL when a 2 mL GPC loop is used) and 4,4'-DDT ,
                     Endrin, and Dieldrin at 0.2 ug/mL  (0.50 ug/mL for a 2 mL
                     loop).  The Aroclor mixture contains 2 ug/mL each of

                                   D-46/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                            Cleanup Procedures

                     Aroclor 1016 and 1260 in methylene chloride (5.0 ug/mL
                   .  when a 2 mL GPC loop is used).

10.1.8.1.4.3.2       Load the first 5 mL sample loop by using a 10 mL syringe
                     containing 8 mL of the pesticide GPC calibration
                     verification solution.  The Aroclor mixture is loaded
                     into loop #2 in the same manner.  Fractions are collected
                     in an auto sequence by using the GPC program established
                     be the UV detector calibration procedure (Section
                     10.1.8.1.3).

10.1.8.1.4.3.3       The collected GPC calibration verification fraction is
                     transferred to a K-D apparatus, and the collection vessel
                     is rinsed with two additional 10 mL portions of methylene
                     chloride to complete the transfer.  The volume of
                     methylene chloride is reduced according to Section
                     10.1.6.1.  After cooling, the solvent is exchanged to
                     hexane according to the instructions in Section 10.1.6.2.
                     The final volume is adjusted to 10 mL, and the sample is
                     analyzed by GC according to the procedure in Section 10.2
                     (usual protocol).  The analysis must be performed on only
                     one of the GC columns used for sample analysis.

10.1.8.1.4.3.4       The pattern of the Aroclor quantitation peaks and the
                     recovery of each single component analyte must be
                     determined for evaluation and reporting purposes.

10.1.8.1.4.4      Technical Acceptance Criteria for GPC Calibration
                  Verification

10.1.8.1.4.4.1       The recovery of  each of the single component analytes
                     must be between  80 - 110 percent.

10.1.8.1.4.4.2       The Aroclor patterns must be the  same as those from the
                     Aroclor 1016 and Aroclor 1260  standards in the initial
                     calibration sequence  (Section  9.2.5.10).

10.1.8.1.4.5      Corrective Action  for GPC Calibration Verification

                  Analysts may  continue to use the  GPC column if  the technical
                  acceptance criteria for  the GPC calibration verification are
                  met.   If the  recoveries  are out of the acceptance window or
                  if changes in the  relative peak heights  of the  patterns of
                  the Aroclor are observed, the columns must be replaced and
                  the GPC  recalibrated according to the instructions in
                  Section  10.1.8.1.3  before proceeding with  any GPC cleanup on
                  samples, including blanks - method and/or  sulfur and MS/MSD.

10.1.8.1.5      Daily UV Calibration  Check (Optional)

                The calibration  of  the GPC may be monitored daily  by use of the
                UV-GPC calibration  solution (Section 7.2.4.3) and  the UV
                detector calibration  procedure  (Section 10.1.8.1.3.3).  The UV
                detector should  be used to monitor the  elution times for the

                                   D-47/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure
Cleanup Procedures

               phthalate, methoxychlor and perylene, in that order.  The pre-
               calibrated GPC program should "dump" > 85 percent of the
               phthalate and should "collect" > 95 percent of the methoxychlor
               and perylene.  Significant changes in elution times of the
               analytes  (e.g., > 0.5 minutes) indicate that the column is out
               of calibration and must be recalibrated or replaced.

10.1.8.1.6     Sample Cleanup by GPC

10.1.8.1.6.1      Introduction to Sample Cleanup by GPC

10.1.8.1.6.1.1   '    It is very important to have consistent laboratory
                     temperatures during an entire GPC run, which could be 24
                     hours or more.  If temperatures are not consistent,
                     retention times will shift, and the dump and collect
                     times determined by the calibration standard no longer
                    •will be appropriate.  The ideal laboratory temperature to
                     prevent out gassing of, the methylene chloride is 22 °C.

10.1.8 ..1-6.1.2       In order to prevent overloading of the GPC column, highly
                     viscous sample extracts must be diluted prior to cleanup.
                     Any sample extract with a viscosity greater than that of
                     1:1 glycerol:water solution must be diluted and loaded
                     into several loops.  Similarly, extracts containing more
                     than 40 mg/mL of non-volatile residue must be diluted and
                     loaded into several loops.  The non-volatile residue may
                     be determined by evaporating a 100 uL aliquot of the
                     extract to dryness in a tared aluminum weighing pan, or
                     other suitable container.  When multiple loops/runs are
                     necessary for ah individual sample, be sure to combine
                     all of the sample eluates collected.

10.1.8.1.6.1.3       Systems using automated injection devices to load the
                     sample on the column must be carefully monitored to
                     assure that the required amount is injected onto the
                     column.  Viscous extracts, or extracts containing a large
                     amounts of non-volatile residue, will cause problems with
                     injecting the- proper amount of sample extract onto the
                     column using automated injection systems.  After the
                     sample extract has been processed, the remaining sample
                     extract in an injection vial must be checked to assure
                     that the proper amount of extract was injected on the
                     column before proceeding with the sample analysis.  If
                     the proper amount of extract was not injected, the sample
                     must be reprepared and the sample extract must be either
                     diluted and loaded into several loops or the sample
                     extract must be injected manually.

10.1.8.1.6.2      Frequency of Sample Cleanup by GPC

                 •GPC cleanup must be performed at least once for each
                  soil/sediment extract, water extracts that contain high
                                   •D-48/PEST                            OLM04 .2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                            Cleanup Procedures

                  molecular weight contaminants that interfere with the
                  analysis of the target analytes and all associated QC
                  (blanks and spikes).   If cleanup procedure is inadequate
                  contact SMO.

10.1.8.1.6.3      Procedure for Sample Cleanup by GPC

10.1.8.1.6.3.1       Particles greater than 5 microns may scratch the valve,
                     which may result in a system leak .and cross contamination
                     of sample extracts in the sample loops.  To avoid such
                     problems, filter the extract through a 5 micron filter
                     disc by attaching a syringe filter assembly containing
                     the filter disc to a 10 mL syringe.  Draw the sample
                     extract through the filter assembly and into the 10 mL
                     syringe.  Disconnect the filter assembly before
                     transferring the sample extract into a small glass
                     container (e.g., a 15 mL culture tube with a Teflon-lined
                     screw cap).   Alternatively, draw the extract into the
                     syringe without the filter assembly.  Attach the filter
                     assembly and force the extract through the filter and
                     into the glass container.  Draw a minimum of 8 mL of
                     extract into a 10 mL syringe.

                     Note:  Some GPC instrument manufacturer' s recommend us'ing
                     a smaller micron size filter disc.  In this instance,
                     follow the manufacturer's recommended operating
                     instructions.

10.1.8.1.6.3.2       INTRODUCTION OF PARTICULATES OR GLASS WOOL INTO THE GPC
                     SWITCHING VALVES MAY REQUIRE FACTORY REPAIR OF THE
                     APPARATUS.

10.1.8.1.6.3.3       Follow  the manufacturer's instructions for operation of
                     the  GPC system being utilized.  Note:  These  instructions
                     were written for a 5 mL GPC injection loop.  A 2 mL
                     injection loop may be used in place of a 5 mL injection
                     loop.   If a 2 mL loop is used, concentrate the 10 mL
                     extract to 4 mL, and then inject 2 mL from the 4 mL.

10.1.8.1.6.3.4       If the  sample is difficult to load, some part of the
                     system may be blocked.  Take appropriate corrective
                     action  following, manufacturer's recommendations.  The
                     problem must be resolved prior to  loading sample
                     extracts.

10.1.8.1.6.3.5       After loading each sample loop, wash the loading port
                     with methylene chloride to minimize cross contamination.
                     Inject  approximately 10 mL of methylene chloride to rinse
                     the  common tubes.

10.1.8.1.6.3.6       After loading all sample  loops, process each  sample using
                     the  collect  and dump cycle times established  in Section
                     10.1.8.1.3.
                                   D-49/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure
Cleanup Procedures

10.1.8.1.6.3.7       Collect each sample in a 250 mL Erlenmeyer flask covered
                     with aluminum foil to reduce solvent evaporation, or
                     directly into a Kuderna-Danish evaporator.  Monitor
                     sample volumes collected.  Changes in sample volumes
                     collected may indicate one or more of the following
                     problems:

                     •  Change in solvent flow rate, caused by channeling in
                        the column or changes in column pressure.

                     •  Increase in column operating pressure due to the
                        absorption of particles or gel fines onto either the
                        guard column or the analytical column gel, if a guard
                        column is not used.

                     •  Leaks in the system or significant variances in room
                        temperature.

10.1.8.1.6.3.8       After the appropriate GPC fraction has been collected for
                     each sample, concentrate the extract as per Section
                     10.1.6.1 and proceed to solvent exchange into hexane as
                     described in Section 10.1.6.2 and Florisil cleanup in
                     10.1.8.2.

10.1.8.1.6.3.9       Any samples that were loaded into two or more loops must
                     be recombined before proceeding with concentration.

10.1.8.2      Florisil Cleanup

10.1.8.2.1     Introduction

               Florisil cartridge cleanup significantly reduces matrix
               interference caused by polar compounds and is required for all
               extracts.  The same volume of the concentrated extract taken
               for Florisil cleanup must, be maintained after Florisil cleanup
               (1 or 2 mL) .

10.1.8.2.2     Florisil Cartridge Performance Check

10.1.8.2.2.1      Summary of Florisil Cartridge Performance Check

                  Every lot number of Florisil cartridges must be tested
                  before they are used for sample cleanup.

10.1.8.2.2.2      Frequency of Florisil Cartridge Performance Check

                  Cartridge performance check must be conducted at least once
                  on each lot of cartridges used for sample cleanup or every 6
                  months, whichever is most frequent.
                                   D-50/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                            Cleanup Procedures

10.1.8.2.2.3      Procedure for Florisil Cartridge Performance Check

                  Add 0.5 mL of 2,4,5-trichlorophenol solution (0.1 ug/mL in
                  acetone, Section 7.2.4.4) and 0.5 mL of Standard Mixture A,
                  midpoint concentration,  (Section 7.2.4.7) to 4 mL of hexane.
                  Reduce the final volume to 0.5 mL using nitrogen (Section
                  10.1.7.2) .   Place the mixture onto the top of a washed
                  Florisil Cartridge, and elute it with 9 mL of hexane/acetone
                  [(90:10)(V/V)].   Use two additional 1 mL hexane rinses to
                  ensure quantitative transfer of standard from the cartridge.
                  Reduce the final volume to 1 mL using nitrogen  (Section
                  10;1.7.2) and analyze the solution by GC/EC using at least
                  one of the GC columns specified for sample analysis.
                  Determine the recovery of each analyte for evaluation and
                  reporting purposes.  Calculate the percent recovery using
                  Equation 12.

                  EQ. 12

                          Percent Recovery = —- x  100



                  Where,

                  Qd =  Quantity determined by  analysis
                  Qa =  Quantity added

10.1.8.2.2.4      Technical Acceptance Criteria for Florisil Cartridge
                  Performance Check

10.1.8.2.2.4.1       The  cartridge performance  check  solution must be analyzed
                     on a GC/EC meeting  the initial calibration  and
                     calibration verification  technical  acceptance criteria.

10.1.8.2.2.4.2       The  lot of Florisil cartridges is acceptable if all
                     pesticides are  recovered  at  80 to 120 percent, if  the
                     recovery of trichlorophenol  is less than 5  percent,  and
                     if no peaks interfering with the target analytes.are
                     detected.

10.1.8.2.2.5      Corrective Action  for  Florisil  Cartridge Performance  Check

                  Any lot of Florisil cartridges  that does not meet the
                  criteria above must be discarded and a new lot, meeting
                  criteria, used for sample cleanup.
                                   D-51/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure
Cleanup Procedures

10.1.8.2.3     Sample Cleanup by Florisil Cartridge

               The required Florisil cartridge size and the final volume of
               the extract after Florisil cleanup are a function of the GC
               autosampler that a laboratory uses.  If the autosampler
               operates reliably with 1 mL of sample extract, then a 500 mg
               cartridge is used and the required final volume is 1 mL.  If
               the autosampler requires more sample, prepare 2 mL of sample
               extract using a 1 g cartridge.  Manual injection requires only
               a 1 mL final extract and a 500 mg cartridge.

10.1.8.2.3.1      Frequency of Sample Cleanup by Florisil Cartridge

                  All sample extracts are required to be cleaned up by the
                  Florisil cartridge technique.

10.1.8.2.3.2      Procedure for Sample Cleanup by Florisil Cartridge

10.1.8.2.3.2.1       Attach the vacuum manifold .to a water aspirator or to a
                     vacuum pump with a trap installed .between the manifold
                     and the vacuum source.  Adjust the vacuum pressure in the
                     manifold to between 5 and 10 pounds of vacuum.

10.1.8.2.3.2.2       Place one Florisil cartridge into the vacuum manifold for
                     each sample extract.

10.1.8.2.3.2.3       Prior to cleanup of samples, the cartridges must be
                     washed with hexane/acetone  (90:10).  This is accomplished
                     by placing the cartridge on the vacuum manifold, by
                     pulling a vacuum, and by passing at least 5 mL of the
                     hexane/acetone solution through the cartridge.  While the
                     cartridges are being washed, adjust the vacuum applied to
                     each cartridge so that the flow rate through each
                     cartridge is approximately equal.  DO NOT ALLOW THE
                     CARTRIDGES TO GO DRY AFTER THEY HAVE BEEN WASHED.

10.1.8.2.3.2.4       After the cartridges on the manifold are washed, the
                     vacuum is released, and a rack containing labeled 10 mL
                     volumetric flasks is placed inside the manifold.  Care
                     must be taken to ensure that the solvent line from each
                     cartridge is placed inside of the appropriate volumetric
                     flask as the manifold top is replaced.

10,. 1. 8 .2 . 3 .2 .5       After the volumetric flasks are in place, the vacuum to
                     the manifold is restored, and a volume of extract equal
                     to the required final volume (1 or 2 mL) from each
                     sample, blank or matrix spike extract is transferred to
                     the top frit of the appropriate Florisil cartridge.  This
                     must equal the final volume after Florisil cleanup.-

10.1.8.2.3.2.6       Because the volumes marked on concentrator tubes are not
                     necessarily accurate at the 1 mL level, the use of a
                     syringe or a volumetric pipet is required to transfer the
                     extract to the cleanup cartridge.

                                   -D-52/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                            Cleanup Procedures

10.1.8.2.3.2.7       The pesticides/Aroclors in the extract concentrates are
                     then eluted through the column with 8 mL of
                     hexane/acetone (90:10) and are collected into the 10 mL
                     volumetric flasks held in the rack inside the vacuum
                     manifold.

10.1.8.2.3.2.8       Transfer the eluate in each volumetric flask to a clean
                     centrifuge tube or 10 mL vial.  Use two additional 1 mL
                     hexane rinses to ensure quantitative transfer of the
                     cartridge eluate.

10.1.8.2.3.2.9       Adjust the extract to the same 1 or 2 mL aliquot volume
                     as was taken for cleanup using either nitrogen blowdown
                     (Section 10.1.7.2) or a micro Snyder column  (Section
                     10.1.7.1).  Measure the final volume with a syringe or by
                     transferring the extract to a volumetric flask.

10.1.8.2.3.2.10      If sulfur cleanup is to be performed, proceed to Section
                     10.1.8.3.  Otherwise, transfer the sample to a GC vial
                     and label the vial.  The extract is ready for GC/EC
                     analysis.

10.1.8.3    Sulfur Cleanup

10.1.8.3.1     Introduction to Sulfur Cleanup

10.1.8.3.1.1      Sulfur contamination will cause a rise in the baseline of a
                  chromatogram and may interfere with the analyses of the
                  later eluting pesticides.  If crystals of sulfur are evident
                  or if the presence of sulfur is suspected, sulfur removal
                  must be performed.   Interference which is due to sulfur is
                  not acceptable.  Sulfur  can be removed by one of two
                  methods, according to laboratory preference.  If the sulfur
                  concentration is such that crystallization occurs in the
                  concentrated extract, centrifuge the extract to  settle the
                  crystals, and remove the sample extract with a disposable
                  pipette, leaving the excess sulfur in the centrifuge tube.
                  Transfer the extract to  a clean centrifuge tube  or  clean
                  concentrator tube before proceeding with further sulfur
                  cleanup.

10.1.8.3.1.2      If only part of a set of samples require sulfur  cleanup,
                  then, a sulfur cleanup blank is required for that part of
                  the set  (Section 12.1.3).

10.1.8.3.2     Frequency of Sulfur Cleanup

               Sulfur removal is required  for all  sample extracts  that contain
               sulfur.
                                   D-53/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors  -- Section 10
Procedure
GC/EC Analysis

10.1.8.3.3     Procedure  for Sulfur Cleanup

10.1.8.3.3.1      Mercury Technique
                  Add one  to  three drops of mercury  to  each hexane extract in
                  a  clean  vial.  Tighten the  top  on  the vial and agitate the
                  sample for  30  seconds.  Filter  or  centrifuge the extract.
                  Pipet the extract  to another vial  and leave all solid
                  precipitate and liquid mercury-  If the mercury appears
                  shiny, proceed to  Section 10.2  and analyze the extract.  If
                  the mercury turns  black, repeat sulfur removal as necessary.
                  The extract transferred to  the  vial still represents the 1.0
                  or 2.0 mL final volume.  CAUTION:  Waste containing mercury
                  should be segregated and disposed  of  properly.  NOTE:
                  Mercury  is  a highly toxic metal and therefore must be used
                  with great  care.   Prior to  using mercury, it is recommended
                  that the analyst become acquainted with proper handling and
                  cleanup  techniques associated with this metal.
 10.1.8.3.3.2       Copper Technique

                   Add approximately 2  g  of  cleaned copper powder  to  the
                   extract in the  centrifuge or concentrator  tube  (2  g will
                   fill the tube to  about .the 0.5  mL mark) .   Mix the  copper and
                   extract for at  least 1 minute on a mechanical shaker.
                   Separate the extract from the copper powder by  drawing off
                   the extract with  a disposable pipet,  and transfer  the
                   extract to a clean vial.   The extract transferred  to the
                   vial still represents  the 1.0 or 2.0 mL final volume.  The
                   separation of the extract from  the copper  powder is
                   necessary to prevent degradation of the pesticides.  If the
                   copper appears  bright, proceed  to Section  10.2  and analyze
                   the extracts.   If the  copper changes color, repeat the
                   sulfur removal  procedure  as necessary.

 10.2   GC/EC Analysis

 10.2.1   Introduction to Sample Analysis by GC/EC

 10.2.1.1    Before samples or required blanks can be analyzed, the instrument
            must  meet the initial calibration and calibration verification
„           technical acceptance  criteria.   Sample analysis  on both  GC columns
            is  required for all samples, blanks,  matrix spikes, and  matrix
            spike duplicates.

 10.2.1.2    Sample extracts, standards,  and blanks must be analyzed  within an
            analytical sequence as  defined  in Section 10.2.2.1, under the same
            instrumental conditions.

 10.2.1.3    Set up the GC/EC system per  the requirements in  Section  9.0.
            Unless ambient temperature on-column  injection is used  (see
            Section 9.1.4),  the injector must be  heated to at least  200 °C.
            The optimized gas chromatographic conditions from Section 9.1 must
            be  used.

                                    D-54/PEST                           OLM04.2

-------
                                   Exhibit D Pesticides'/Aroclors -- Section 10
                                                                     Procedure
                                                                GC/EC Analysis

10.2.2   Procedure for Sample Analysis by GC/EC

         The injection must be made on-column by using either automatic or
         manual injection.  If autoinjectors are used, 1 uL injection volumes
         may be used.   Manual injections shall use at least 2 uL injection
         volumes.   The same injection volume must be used for all standards,
         samples,  MS/MSD, and blanks associated with the same initial
         calibration.   If a single injection is used for two GC columns
         attached to a single injection port, it may be necessary to use an
         injection volume greater than 2 uL.  However, the same injection
         volume must be used for all analyses.

10.2.2.1    Analytical Sequence

            All acceptable samples must be analyzed within a valid analysis
            sequence as given below.
 Time
 0 hr.
Infection #
1-15


16


17


18
Material Injected
First 15 steps of the initial
calibration
Instrument Blank at end of
initial calibration
PEM at end of initial
calibration
First sample
Subsequent samples
 12 hr.
                        1st injection past
                        12 hr.
                        2nd and 3rd injections
                        past 12 hr.
                         Last Sample
                         Instrument blank


                         Individual Standard Mixtures
                         A and B
                         Sample
                                                 Subsequent samples
 Another  12 hr.
                        1st injection past
                        12 hr.
                        2nd injection past
                        12 hr.
                         Last Sample
                         Instrument blank

                         PEM

                         Sample
                                                 Subsequent samples
                                   D-55/PEST
                                                OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 10
Procedure
GC/EC Analysis

 Time                   Injection #              Material Injected


 Another 12 hr.       •  .                        Last Sample
                        1st  injection past       Instrument blank
                        12 hr;
                        2nd  and 3rd  injections  Individual Standard Mixtures
                        past 12  hr.              A and B
                                                Sample


                        etc.

10.2.2.1.1     NOTE:  The first  12 hours are counted from injection #16  (the
               instrument blank  at the end of the initial calibration
               sequence), not from injection #1.  Samples may be injected
               until 12 hours have elapsed.  All subsequent 12-hour periods
               are timed from the injection of the instrument blank that
               brackets the front end of the samples.  Because the 12-hour
               time period  is timed  from injection of the instrument blank
               until the injection of the last sample, each 12-hour period may
               be separated by the length of one chromatographic run, that of
               the analysis of the last sample.  While the 12-hour period may
               not be exceeded,  the  laboratory may run instrument blanks and
               standards more frequently for instance, to accommodate staff
               working  8-hour shifts.

10.2.2.1.2     After the initial calibration, the analysis sequence may
               continue as  long  as acceptable instrument blanks, PEMs and
               Individual Standard Mixtures A and B are analyzed at the
               required frequency.   This analysis sequence shows only the
               minimum  required  blanks and standards.  More blanks and
               standards may be  run  at the discretion of the Contractor:
               these must also satisfy the criteria presented in Section 9 in
               order to continue the run secruence.

10.2.2.1.3     An analysis  sequence  must also include all required matrix
               spike/matrix spike duplicate and method  (and/or sulfur) blank
               analyses, but the Contractor may decide at what point in the
               sequence they are to  be analyzed.

10.2.2.1.4     The requirements  for  the analysis sequence apply to both GC
               columns  and  for all instruments used for these analyses.

10.2.3   Sample Dilutions

10.2.3.1    All samples must be  analyzed at the most concentrated level that
            is consistent with achieving satisfactory chromatography  (defined
            in Section  11.3) .
                                   D-56/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 10
                                                                     Procedure
                                                                GC/EC Analysis

10.2.3.2    If the response of any single component pesticide is greater than
            the response of that analyte in the initial calibration high point
            standard, then the extract must be diluted to have the response of
            that analyte between the initial calibration low point and high
            point standards.

10.2.3.3    If the response of the largest peak in a multicomponent analyte is
            greater than the most intense single component analyte response in
            the initial calibration high point standard, then the sample must
            be diluted to have the response of the largest peak in the
            multicomponent analyte between the responses of the initial
            calibration midpoint and high point standards of that single
            component pesticide.

10.2.3.4    If dilution is employed solely to bring a peak within the
            calibration range or to get a multicomponent pattern on scale, the
            results for both the more and the less concentrated extracts must
            be reported.  The resulting changes in quantitation limits and
            surrogate recovery must be reported also for the diluted samples.

10.2.3.5    If the Contractor has reason to believe that diluting the final
            extracts will be necessary, a less diluted run may still be
            required.  If an acceptable chromatogram  (as defined in Section
            11.3) is achieved with the diluted extract, then:

            • If the dilution factor is greater than 10, an additional
               extract 10 times more concentrated than  the diluted sample
               extract must be injected and reported with the sample data.

            • If the dilution factor is less than or equal to 10, then an
               undiluted sample extract must be injected and reported with the
               sample data.

            If the analysis of the most concentrated extract does not meet the
            requirement for dilution in Section 11.3.5, then the analysis is
            at no additional cost to the Agency.

10.2.3.6    When diluted, the chromatographic data for  the single component
            pesticides must be able to be reported at greater than 10.0
            percent  of full scale but less than 100.0 percent of full scale.

10.2.3.7    When diluted, multicomponent analytes must be able to be reported
            at greater than 25.0 percent of full scale but less than 100.0
            percent  of full scale.

10.2.3.8    If a chromatogram is replotted electronically to meet these
            requirements, the scaling factor used must  be displayed on the
            chromatogram.  If the chromatogram of any sample needs to be
            replotted electronically to meet these requirements both the
            initial  chromatogram and the replotted chromatogram must be
            submitted in the data package.
                                   D-57/PEST                            OLM04.2

-------
  Exhibit D Pesticicies/Aroclors -- Section 10
  Procedure
  GC/EC Analysis

  10.2.3.9    Samples with analytes detected at a level greater than the high
              calibration point must be diluted until the response is within the
              linear range established during calibration or to a maximum of
              1:100,000.

  10.2.3.10   If the response is still above the high calibration point after
              the dilution of 1:100,000, the Contractor shall contact SMO
              immediately.

  10.2.3.11   Use the results of the original analysis to determine the
              approximate dilution factor required to get the largest analyte
              peak within the initial calibration range.

  10.2.3.12   The dilution factor chosen should keep the response of the largest
              peak for a target compound in the upper half of the initial
              calibration range of the instrument.

  10.2.3.13   Sample dilutions must be made quantitatively.  Dilute the sample
              extract with hexane.

|  10.2.3.14   Do not submit data for more than two analyses, i.e., from the
              original sample extract and one dilution, or, from the most
              concentrated dilution analyzed and one further dilution.  This
              statement does not refer .to reanalyses required due to failed
              technical acceptance criteria.
                                     D-58/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 11
                                                Data Analysis and Calculations
                                                    Qualitative Identification

11.0  DATA ANALYSIS AND CALCULATIONS

11.1  Qualitative Identification

11.1.1   Identification of Target Compounds

11.1.1.1    The laboratory will identify and quantitate analyte peaks based on
            the RT windows and the calibration factors of the midpoint
            standard (single component pesticides)  established during the
            initial calibration sequence.

11.1.1.2    Analytes are identified when peaks are observed in the RT window
            for the analyte on both GC columns.

11.1.1.3    A set of three to five major peaks is selected for each
            multicomponent analyte.  The retention time window for each peak
            is determined from the initial calibration analysis.
            Identification of a multicomponent analyte in the sample is based
            on pattern recognition in conjunction with the elution of three to
            five sample peaks within the retention time windows of the
            corresponding peaks of the standard on both GC columns.
            Calibration factors used to quantitate toxaphene and the Aroclors
            are based on the single-point calibration standard analyzed during
            the initial calibration.  The number of potential quantitation
            peaks is listed in Table 2.

11.1.1.4    When any multicomponent analyte is detected in a sample, a
            standard must be run within 72 hours of the analyte's detection
             (from time of injection), and within a valid 12-hour sequence.

11.1.1.5    The choice of the peaks used for multicomponent analyte
            identification and the recognition of those peaks may be
            complicated by the environmental alteration of the  toxaphene or
            Aroclors, and by the presence of co-eluting analytes or matrix
            interferences, or both.  Because of the alteration  of these
            materials in the environment, multicomponent analytes in samples
            may give patterns similar to, but not identical with, those of the
            standards.

11.1.1.6    If more than one multicomponent analyte is observed in a sample,
            the Contractor must choose different peaks to quantitate each
            multicomponent analyte.  A peak common to both analytes present in
            the sample must not be used to quantitate either compound.

11.1.2    GC/MS Confirmation of Pesticides and Aroclors

11.1.2.1    Any pesticide or Aroclor analyte listed in Exhibit  C for which a
            concentration is reported  from a GC/EC analysis must have the
            identification confirmed by GC/MS if the concentration is
            sufficient for that purpose.  The following paragraphs are to be
            used as guidance in performing GC/MS confirmation.  If the
            Contractor fails to perform GC/MS confirmation as appropriate, the
            Agency may require reanalysis of any affected samples  at no
            additional cost to the Agency.
                                   D-59/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 11
Data Analysis and Calculations
Qualitative Identification

11.1.2.2    The GC/MS confirmation may be accomplished by one of three general
            means:

            •   Examination of the semivolatile GC/MS library search results
                (i.e.,TIC data), or

            •   A second analysis of the semivolatile extract, or

            •   Analysis of the pesticide/Aroclor extract, following any
                solvent exchange and concentration steps that may be
                necessary.

11.1.2.3    The semivolatile GC/MS analysis procedures outlined in Exhibit D
            SVOA are based on the injection into the instrument of
            approximately 20 ng of a target compound in a 2 uL volume.  The
            semivolatile CRQL values in Exhibit C are based on the sample
            concentration that corresponds to an extract concentration of 10
            ng/uL of target analyte.  However, these are quantitation limits,
            and the detection of analytes and generation of reproducible mass
            spectra will routinely be possible at levels 3-10 times lower.
            The sample concentration corresponding to 10 ng/uL in extract will
            depend on the sample matrix.

11.1.2.3.1     For water samples, 20 ng/2pL corresponds to a sample
               concentration of 10 ug/L.

11.1.2.3.2     For soil/sediment samples prepared according to the
               semivolatile low level soil/sediment method (i.e., 30 g of
               soil/sediment), the corresponding sample concentration is 330
               ug/Kg.

11.1.2.3.3     For soil/sediment samples prepared according to the
               semivolatile medium'level soil/sediment method (i.e., 1 g of
               soil/sediment), the corresponding sample.concentration is
               10,000 ug/Kg.

11.1.2.3.4     Therefore, based on the values given above, any pesticide
               sample in which compound concentration in the sample extract is
               greater than or equal to 10 ng/uL for single component
               pesticides, 50 ng/uL for Aroclors, and 125 ng/uL for Toxaphene
               should enable the laboratory to confirm the pesticide/Aroclor
               by GC/MS analysis of the semivolatile extract.

11.1.2.4    In order to confirm the identification of the target
            pesticide/Aroclor, the laboratory must also analyze a reference
            standard for the analyte.  In order to demonstrate the ability of
            the GC/MS system to identify the analyte in question, the
            concentration of the standard should be 10 ng/uL for single
            component pesticides, 50 ng/uL for Aroclors, and 125 ng/uL for
            Toxaphene.

11.1.2.5    To facilitate the confirmation of the single component pesticide
            analytes from the semivolatile library search data,  the laboratory
            may wish to include these analytes in the semivolatile continuing
            calibration standard at a concentration of 10 ng/uL or less.  Do


                                   D-60/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 11
                                                Data Analysis and Calculations
                                                    Qualitative Identification

            not include the Aroclors and toxaphene mixture in the semivolatile
            initial and continuing calibration standard.   If added to this
            GC/MS standard, the response factors,  retention times, etc.  For
            these analytes would be reported on the GC/MS quantitation report,
            but not on the GC/MS calibration data reporting forms.  As only a
            single concentration of each analyte would be analyzed, no
            linearity (%RSD) or percent difference criteria would be applied
            to the response factors for these additional analytes.

11.1.2.6    The laboratory is advised that library search results from the
            NIST/EPA/NIH  (May 1992 release or later) and Wiley  (1991 release
            or later) mass spectral library will not likely list the name of
            the pesticide/Aroclor analyte as it appears in this SOW, hence,
            the mass spectral interpretation specialist is advised to compare
            the CAS Registry numbers for the pesticides/Aroclors to those from
            the library search routine.

11.1.2.7    If the analyte cannot be confirmed from the semivolatile library
            search data for the original semivolatile GC/MS analysis, the
            laboratory may analyze another aliquot of the semivolatile sample
            extract after further concentration of the aliquot.  This second
            aliquot must either be analyzed as part of a routine  semivolatile
            GC/MS analysis, including instrument performance checks  (DFTPP),
            calibration standards containing the pesticides/Aroclors as
            described in Section 11.1.2.5, or it must be analyzed along with
            separate reference standards for the analytes to be confirmed.

11.1.2.8    If the analyte cannot be confirmed by either the procedures in
            Sections 11.1.2.5 or 11.1.2.7, then an aliquot of the extract
            prepared for the GC/EC analysis must be analyzed by GC/MS,
            following any necessary solvent exchange and concentration steps.
            As in Section  11.1.2.4, analysis of a reference standard is
            required if the GC/MS continuing calibration standard does not
            contain the analyte to be confirmed.

11.1.2.9    Regardless of which of the  three approaches above is  used  for
            GC/MS confirmation, the appropriate blank must also be analyzed by
            GC/MS to demonstrate that the presence of the analyte was not the
            result of laboratory contamination.  If the confirmation is based
            on the analysis of the semivolatile extract, then the  semivolatile
            method blank  extracted with the sample must also be analyzed.   If
            the confirmation is based on the analysis of the extract prepared
            for the GC/EC  analysis, then the pesticide method blank extracted
            with the sample must be analyzed.

11.1.2.10   If the identification of the analyte cannot be confirmed by any of
            the GC/MS procedures above  and the concentration calculated from
            the GC/EC analysis is greater than or equal to the  concentration  -
            of the reference standard analyzed by GC/MS, then report the  .
            analyte  as undetected, adjust the sample quantitation limit  (the
            value associated with the  "U" qualifier) to a sample  concentration
            equivalent to the  concentration of the  GC/MS reference  standard,
            and qualify the results on  Form I with  one of the laboratory
            defined  qualifiers  ("X,"  "Y," or  "Z").   In this  instance,  define
                                   D-61/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 11
Data Analysis and Calculations
Calculations

            the qualifier explicitly in the SDG Narrative, and describe the
            steps taken to confirm the analyte in the SDG Narrative.

11.1.2.11   For GC/MS confirmation of single component analytes, the required
            deliverables are copies of the library search results '(best TIC
            matches) or analyte spectrum and the spectrum of the reference
            standard.  For multicomponent analytes, spectra of three
            characteristic peaks are required for both the sample component
            and the reference standard.

11.1.2.12   The purpose of the GC/MS analysis for the single component
            pesticides is for identification.  The purpose of the GC/MS
            analysis for the multicomponent analytes is to confirm the
            presence of chlorinated biphenyls in Aroclor and the presence of
            chlorinated camphenes in Toxaphene.  The GC/MS analytical results
            for the pesticides/Aroclors shall not be used for quantitation and
            the GC/MS results shall not be reported on Form I and Form X.  The
            exception noted in Section 11.1.2.10 applies only to analytes that
            cannot be confirmed above the reference standard concentration.

11.2  Calculations

11.2.1   Target Compounds

         The concentrations of the single component pesticides and surrogates
         are calculated separately for both GC columns by using the following
         equations.

11.2.1.1    Water

            EQ. 13
                                         (Ax) (Vt) (Df) (GPC)
                    Concentration ug/L = 	
                                            (CF) (Vo) (V,)
            Where,

            AX  =  Area of  the peak  for  the  compound  to be measured.
            CF  =  Calibration factor  from the  initial calibration  for  the
                  midpoint concentration external  standard  (area per ng).
            V0  =  Volume of water extracted in milliliters  (mL).
            Vi  =  Volume of extract injected in microliters  (uL).   (If a
                  single injection  is made  onto two  columns, use one half the
                  volume in the  syringe as  the volume injected  onto each
                  column.)

            Vt  =  Volume of the  concentrated extract in microliters  (uL).   (If
                  GPC is not performed, then Vt = 10,000 uL.  If GPC is
                  performed, then Vc = 5,000 uL.)

            Df  =  Dilution factor.  The dilution factor for  analysis of water
                  samples  by this method is defined  as follows:
                                   D-62/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 11
                                                Data Analysis and Calculations
                                                                  Calculations
         uL most cone, extract used to make dilution + uL clean solvent
                  uL most cone,  extract used to make dilution
            If no dilution is performed, Df = 1.0.

            GPC = GPC factor.  (If no GPC is performed, GPC = 1.  If GPC is
            performed, then GPC =2.0
11.2.1.2    Soil/Sediment

            EQ. 14
                                                   (Ax) (Vt) (Df) (GPC)
           Concentration ug/Kg (Dry weight basis)  =
                                                    (CF) (V,) (Wa) (D)
            Where,

            AX and CF are as given for water,  above.
            Vt =  5,000 uL.
            Vi =  Volume of extract injected in microliters  (uL).   (If  a
                  single injection is made onto two columns, use one half the
                  volume in the syringe as the volume injected onto each
                  column.)
                               _  100 - % moisture
                                        Too
            Ws =  Weight of sample extracted  in grams  (g)
            Df =  Dilution factor.  The dilution  factor  for analysis  of
                  soil/sediment samples by  this method is defined  as  follows:

          uL most  cone. Extract used to make dilution + uL  clean solvent
                   uL most cone, extract used to make dilution
                   If no  dilution  is performed,  Df  =  1.0.

            GPC  =  GPC  factor  =  2.

11.2.1.2.1  "   The GPC factor is  used  to account  for the  amount  of  extract
               that is not  recovered from the mandatory use  of GPC  cleanup.
               Concentrating  the  extract collected after  GPC to  5.0 mL rather
               than 10.0 mL for water  samples not  subjected  to GPC  maintains
               the sensitivity  of the  soil/sediment  method comparable to that
               of  the  water method, but  correction of the numerical results  is
               still required.
                                   D-63/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 11
Data Analysis and Calculations
Calculations

11.2.1.2.2     Note that the calibration factors used for the quantitation of
               the single component pesticides are the calibration factors
               from the midpoint concentration on standard in the most recent
               initial calibration.

11.2.1.2.3     Because of the likelihood that compounds co-eluting with the
               target compounds will cause positive interferences and increase
               the concentration determined by the method, the lower of the
               two concentrations calculated for each single component
               pesticide is reported on Form I.  In addition, the
               concentrations calculated for. both the GC columns are reported
               on Form X, along with a percent difference comparing the two
               concentrations.  The percent difference is calculated according
               to Equation 15.
               EQ. 15
                                 Cone.. - ConcT
                            %D =	 x 100
               Where,                Conc
                                         L
               ConcH   =  The higher of the two concentrations for the target
                          compound in question.
               ConcL   =  The lower of the two concentrations for the target
                          compound in question

11.2.1.2.4     Note that using this equation will result in percent  difference
               values that are always positive.  The value will also be
               greater  than  a value calculated using the higher concentration
               in the denominator; however, given the likelihood of  a positive
               interference  raising the concentration determined on  one GC
               column,  this  is a  conservative  approach to comparing  the two
               concentrations.

11.2.1.2.5     The quantitative determination  of Toxaphene or Aroclors is
               somewhat different from that of single component pesticides.
               Quantitation  of peaks within the detector linear range CRQL to
               > 16 times CRQL is based on a single calibration point assuming
               linear detector response.  Alternatively, a linear calibration
               range may be  established during a run sequence by a three-point
               calibration curve  for any multicomponent analyte.  If the
               concentration is calculated to  be 106 times the CRQL,  the
               Contractor shall contact SMO immediately.

11.2.1.2.6     The quantitation of toxaphene or Aroclors must be accomplished
               by comparing  the heights or the areas of each of the  three to
               five major peaks of the multicomponent analyte in the sample
               with the calibration factor for the same peaks established
               during the initial calibration  sequence.  The concentration of
               multicomponent analytes is calculated by using Equations 13 and
               14, where f^ is the area for each of the major peaks of the
               multicomponent analyte.  The concentration of each peak is
               determined and then a mean concentration for the three to five
               major peaks is determined on each column.
                                   D-64/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors  -- Section  11
                                                Data Analysis  and Calculations
                                                                  Calculations

11.2.1.2.7     The reporting requirements for Toxaphene and the Aroclors  are
               similar to those for the single component  analytes, except that
               the lower mean concentration  (from three to five peaks)  is
               reported on Form I, and the two mean concentrations reported  on
               Form X.  The two mean concentrations are compared  by
               calculating the percent difference using Equation  15.

11.2.2   CRQL Calculation

         If the adjusted CRQL is  less than the CRQL listed in  Exhibit  C
          (Pesticides), report the CRQL in Exhibit C  (Pesticides).

11.2.2.1    Water Samples

            EQ. 16


                     Adjusted _ Contract   (Vx) (Vt) (Vy) (Df).
                       CRQL   '    CRQL   X    (v_)  (v_) (V.)
            Where,

            Vt, Df, V0,  and V4 are as given  in Equation  13.
            Vx  =   Contract sample volume (1000  mL).
            Vy  =   Contract injection volume (1  uL or 2  uL).
            Vc  =   Contract concentrated extract volume  (10,000 uL if GPC was
                   not performed and 5,000 uL if GPC  was performed).

11.2.2.2    Soil/Sediment  Samples

            .EQ. 17
                     Adjusted . Contract    (Wx) (Ve) (Vy) (Df)
                       CRQL       CRQL   X   (wj (YC) (VA) (D)
             Where,
             Vt, Df, Ws, V±  and D are as given in Equation 14.
             Wx  =   Contract sample weight (30 g).
             Vy  =   Contract injection volume (1 uL or 2 uL).
             Vc  =   Contract concentrated extract volume  (GPC is required:
                   5,000 uL).

 11.2.3    Surrogate Recoveries

 11.2.3.1    The concentrations of the surrogates are calculated separately for
             each GC column in a similar manner as the other analytes, using
             Equations 13 and 14.  Use the calibration factors from the
             midpoint concentration of Individual Standard Mixture A from the
             initial calibration.  The recoveries of the surrogates calculated
             for each GC column according to Equation 12, repeated below.
                                    D-65/PEST                            OLM04.2

-------
Exhibit D Pesticicies/Aroclors -- Section 11
Data Analysis and Calculations
Technical Acceptance Criteria for Sample Analysis
            EQ. 12
            Where,
                          Percent Recovery = —- x 100
                                             >«a
            Qd =  Quantity determined by analysis
            Qa =  Quantity added

11.2.3.2    The advisory limits for the recovery of the surrogates are 30-150
            percent for both surrogate compounds.

11.2.3.3    As these limits are only advisory, no further action is required
            by the laboratory; however, frequent failures to meet the limits
            for surrogate recovery warrant investigation by the laboratory,
            and may result in questions from the Agency.  Surrogate recovery
            data from both GC columns are reported  (see Exhibit. B).

11.3  Technical Acceptance Criteria for Sample Analysis

      The requirements below apply independently to each GC column and to all
      instruments used for these analyses.  Quantitation must be performed on
      each GC column.

11.3.1   Samples must be analyzed under the GC/EC operating conditions in
         Section 9.  The instrument must have met all initial calibration,
         calibration verification, and blank technical acceptance criteria.
         Samples must be cleaned-up, when required, on a GPC meeting the
         technical acceptance criteria for GPC calibration and GPC calibration
         checks.  Samples must be cleaned-up using florisil meeting the
         technical acceptance criteria for florisil.  Sample data must be
         bracketed at 12-hour intervals  (or less) by acceptable analyses of
         instrument blanks, PEMs and Individual Standard Mixtures A and B, as.
         described in Section 10.2.2.1.

11.3.2   The samples must be extracted and analyzed within the contract
         required holding times.

11.3.3   The samples must have an associated method blank meeting the
         technical acceptance criteria for method blanks.  When sulfur cleanup
         blanks are required, the samples must have associated with it a
         sulfur cleanup blank meeting the technical acceptance criteria for
         sulfur cleanup blanks.

11.3.4   The retention time for each of the surrogates must be within the
         retention time window as calculated in Section 9 for both GC columns.

11.3.5   No target analyte concentrations may exceed the upper limit of the
         initial calibration or else the extract must be diluted and
         reanalyzed.

11.3.6   A standard for any identified multicomponent analyte must be analyzed
         during a valid analytical sequence on the same instrument and column
         within 72 hours of its detection in a sample.


                                   D-66/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 11
                                                Data Analysis and Calculations
                                         Corrective Action for Sample Analysis


11.3.7   The identification of single component pesticides by gas
         chromatographic methods is based primarily on retention time data.
         The retention time of the apex of a peak can be verified only from an
         on-scale chromatogram.  The identification of multicomponent analytes
         is based primarily on recognition of patterns of retention times
         displayed on a chromatogram.  Therefore, the following requirements
         apply to all data presented for single component and multicomponent
         analytes.

11.3.7.1    When no analytes are identified in a sample, the chromatograms
            from the analyses of the sample extract must use the same scaling
            factor as was used for the low point standard of the initial
            calibration associated with those analyses.

11.3.7.2    Chromatograms must display single component pesticides detected in
            the sample at less than full scale.

11.3.7.3    Chromatograms must display the largest peak of any multicomponent
            analyte detected in the sample at-less than full scale.

11.3.7.4    If an extract must be diluted, chromatograms must display single
            component pesticides between 10 and 100 percent of full scale.

11.3.7.5    If an extract must be diluted, chromatograms must display the
            peaks chosen for quantitation of multicomponent analytes between
            25 and 100 percent of full scale.

11.3.7.6    For any sample or blank, the baseline of the chromatogram must
            return to below 50 percent of full scale before the elution time
            of alpha-BHC, and return to below 25 percent of full scale after
            the elution time of alpha-BHC -and before the elution time of
            decachlorobiphenyl.

11.3.7.7    If a chromatogram is replotted electronically to meet  these
            requirements, the scaling factor used must be displayed on the
            chromatogram.

11.3.7.8    If the chromatogram of any sample needs  to be replotted
            electronically to meet these  requirements, both the initial
            chromatogram and the replotted chromatogram must be submitted  in
            the data package.

11.4   Corrective Action for Sample Analysis

11.4.1   Sample analysis technical acceptance  criteria MUST be met before  data
          are  reported.  Samples contaminated  from  laboratory sources or
          associated with a contaminated method blank or sulfur  cleanup blank
          will require  re-extraction  and reanalysis  at no additional  cost to
          the  Agency.   Any samples  analyzed that  do  not meet  the technical
          acceptance criteria will  require re-extraction and/or  reanalysis  at
          no additional cost  to the Agency.

11.4.2   If the sample analysis technical acceptance criteria are  not met,
          check calculations, surrogate solutions,  and  instrument performance.


                                   D-67/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 11
Data Analysis and Calculations
Corrective Action for Sample Analysis

         It may be necessary to recalibrate the instrument or take other
         corrective action procedures to meet the technical acceptance
         criteria, in which case, the affected samples must be reanalyzed at
         no additional cost to the Agency after the corrective action.

11.4.3   The extract from samples which were cleaned-up by GPC using an
         automated injection system and have surrogate recoveries outside the
         lower advisory surrogate acceptance limits must be.checked to assure
         that the proper amount was injected on the GPC column.  If
         insufficient volume was injected, the sample must be reprepared and
         reanalyzed at no additional cost to the Agency.

11.4.4   If sample chromatograms have a high baseline or interfering peaks,
         inspect the system to determine the cause of the problem (e.g.,
         carryover, column bleed, dirty BCD, contaminated gases, leaking
         septum, etc.).  After correcting the problem, analyze an instrument
         blank to demonstrate that the system is functioning properly.
         Reanalyze the sample extracts.  If the problem with the samples still
         exists, then those samples must be re-extracted and reanalyzed.
         Samples which cannot be made to meet the given specifications after
         one re-extraction and three-step cleanup  (GPC, Florisil, and sulfur
         cleanups) are reported in the SDG Narrative and do not require
         further analysis.
                                   D-68/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 12
                                                               Quality Control
                                                                Blank Analyses

12.0  QUALITY CONTROL

12.1  Blank Analyses

12.1.1   Introduction

         There are two types of blanks always required by this method:  the
         method blank and the instrument blank.  A separate sulfur cleanup
         blank may be required if all samples associated with a given method
         blank are not subjected to sulfur cleanup.  Samples that are
         associated with a sulfur cleanup blank are also associated with the
         method blank with which they were extracted.  Both the method and
         sulfur cleanup blanks must meet the respective acceptance criteria
         for the sample analysis acceptance criteria to be met.

12.1.2   Method Blanks

12.1.2.1    Summary of Method Blanks

            A method blank is a volume of a clean reference matrix  (reagent
            water for water samples, or purified sodium sulfate for
            soil/sediment samples) that is carried through the entire
            analytical procedure.  The volume or weight of the reference
            matrix must be approximately equal to the volume or weight of
            samples associated with the blank.  The purpose of a method blank
            is to determine the  levels of contamination associated with the
            processing and analysis of samples.

12.1.2.2    Frequency of Method  Blanks

            A method blank must  be extracted each  time samples are  extracted.
            The number of samples extracted with each method blank  shall not
            exceed 20 field samples  (excluding matrix spikes/matrix spike
            duplicates and PE samples).   In addition, a method blank shall:

            •  Be extracted by the same procedure  used to extract  samples.

            •  Be analyzed on each GC/EC  system used to analyze associated
               samples.

12.1.2.3    Procedure for Method Blank Preparation

12.1.2.3.1     For pesticide/Aroclor analyses, a method blank for  water
               samples consists  of a 1 L  volume of reagent water spiked with
               1.0 mL of the surrogate spiking solution  (Section 7.2.4.1).
               For soil/sediment samples,  the method blank consists of 30  g of
               sodium sulfate spiked with 2.0 mL of  the surrogate  spiking
               solution.

12.1.2.3.2     Extract,  concentrate, and  analyze method blanks  according to
               Section 10.

12.1.2.3.3     Calculate method  blank  results  according to Section 11.

12.1.2.4    Technical Acceptance Criteria for  Method Blanks


                                   D-69/PEST                            OLM04.2

-------
Exhibit D Pesticicies/Aroclors -- Section 12
Quality Control
Blank Analyses

12.1.2.4.1     The requirements below apply independently to each GC column
               and to all instruments used for these analyses.  Quantitation
               must be performed on both GC columns.

12.1.2.4.2     All method blanks must be prepared and analyzed at the
               frequency described in Section 12.1.2.2, using the procedure
               above and in Section 10 on a GC/EC system meeting the initial
               calibration and calibration verification technical acceptance
               criteria.  Method blanks must undergo cleanup, when required,
               on a GPC meeting the technical acceptance criteria for GPC
               calibration and GPC calibration checks.  Method blanks must be
               cleaned-up using Florisil meeting the technical acceptance
               criteria for Florisil.  Method blanks must be bracketed at 12-
               hour intervals  (or less) by acceptable analyses of instrument
               blanks, PEMS, and individual standard mixtures A and B as
               described in Section 10.2.2.1.
                                                      I
12.1.2.4.3     The concentration of the target compounds  (Exhibit C
                (Pesticides)) in the method blank must be less than the CRQL
               for each target compound.

12.1.2.4.4     The method-blank must meet all sample technical acceptance
               criteria in Sections 11.3.4 to 11.3.7.

12.1.2.4.5     Surrogate recoveries must fall within the acceptance windows of
               30-150%.  In the case of the method blank(s), these limits are
               not advisory.

12.1.2.5    Corrective Action for Method Blanks

12.1.2.5.1     If a method blank does not meet the technical acceptance
               criteria, the Contractor must consider the system to be out of
               control.

12.1.2.5.2     If contamination is a problem, then the source of the
               contamination must be investigated and appropriate corrective
               measures must be taken and documented before further sample.
               analysis proceeds.  It is the Contractor's responsibility to
               ensure that method interferences caused by contaminants in
               solvents, reagents, glassware, and sample storage and sample
               processing hardware that lead to discrete artifacts and/or
               elevated baselines be investigated and appropriate corrective
               actions be taken and documented before further sample analysis.
               All samples associated with a contaminated method blank must be
               re-extracted/reanalyzed at no additional cost to the Agency.

12.1.2.5.3     If the surrogate recoveries in the method blank do not meet the
               acceptance criteria listed in Section 12.1.2.4.5, first
               reanalyze the method blank.  If surrogate recoveries do not
               meet the acceptance criteria after reanalysis, the method blank
               and all samples associated with that method blank must be
               re-extracted and reanalyzed at no additional cost to the
               Agency.

12.1.2.5.4     If the method blank failed to meet the criteria listed in


                                   D-70/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors --  Section 12
                                                               Quality Control
                                                                Blank Analyses

               Sections 12.1.2.4.2 and 12.1.2.4.4,  then there is an instrument
               problem.  Correct the instrument problem and reanalyze the
               method blank.

12.1.3      Sulfur Cleanup Blanks

12.1.3.1    Summary of Sulfur Cleanup Blanks

            The sulfur cleanup blank is a modified form of the method blank.
            The sulfur cleanup blank is hexane spiked with the surrogates and
            carried through the sulfur cleanup and analysis procedures.   The
            purpose of the sulfur cleanup blank is to determine the levels of
            contamination associated with the separate sulfur cleanup steps.

12.1.3.2    Frequency of Sulfur Cleanup Blanks

            The sulfur cleanup blank is prepared separately when only part of
            a set of samples extracted together requires sulfur removal.   A
            method blank is associated with the entire set of samples.  The
            sulfur cleanup blank is associated with the part of the set which
            required sulfur cleanup.  If all the samples associated with a
            given method blank are subjected to sulfur cleanup, then the
            method blank must be subjected to sulfur cleanup, and no separate
            sulfur cleanup blank is required.

12.1.3.3    Procedure  for Sulfur Cleanup Blank

12.1.3.3.1     The concentrated volume of the blank must be the same as the
               final volume of the samples associated with the blank.  The
               sulfur  blank must also contain the surrogates at the same
               concentrations as the sample extracts  (assuming 100.0 percent
               recovery).  Therefore, add 0.1 mL of the surrogate solution to
               0.9 mL  of hexane in a clean vial, or for a sulfur blank with a
               final volume of 2 mL, add 0.2 mL of the surrogate solution to
               1.8 mL  of hexane in a clean vial.

12.1.3.3.2     Proceed with the sulfur removal  (Section 10.1.8.3.3.1 or
               10.1.8.3.3.2) using the same technique  (mercury or copper) as
               the samples  associated with the blank.

12.1.3.3.3     Analyze the  sulfur cleanup blank according to Section  10.2.
               Assuming that the material in the sulfur cleanup blank resulted
               from  the extraction of a 1 L water sample, calculate the
               concentration of each analyte using the equation in Section
               11.2.1.1.  Compare the results  to the  CRQL values for  water
               samples in Exhibit C  (Pesticides).

12.1.3.4    Technical  Acceptance  Criteria For  Sulfur  Cleanup Blanks

12.1.3.4.1     The requirements below apply independently to each GC  column
               and to  all instruments used  for these  analyses.  Quantitation
               must  be performed  on both  GC columns.  .
                                   D-71/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 12
Quality Control
Blank Analyses

12.1.3.4.2     All sulfur cleanup blanks must be prepared and analyzed at the
               frequency described in Section 12.1.3.2 using the procedure
               referenced in Section 12.1.3.3 on a GC/EC system meeting the
               initial calibration and calibration verification technical
               acceptance criteria.

12.1.3.4.3     Sulfur cleanup blanks must be bracketed at 12-hour intervals
               (or less) by acceptable analyses of instrument blanks, PEMs,
               and Individual Standard Mixtures A and 8,  as described in
               Section 10.2.2.1.

12.1.3.4.4     The concentration of the target compounds (Exhibit C
               (Pesticides)) in the sulfur cleanup blank must be less than the
               CRQL for each target compound.

12.1.3.4.5     The sulfur cleanup blank must meet all sample technical
               acceptance criteria in Sections 11.3.4 to 11.3.7.

12.1.3.4.6     Surrogate recoveries must fall within the acceptance windows of
               30-150%.  In the case of the sulfur cleanup blank, -these limits
               are not advisory.

12.1.3.5    Corrective Action for Sulfur Cleanup Blanks

12.1.3.5.1     If a sulfur cleanup blank does not meet the technical
               acceptance criteria, the Contractor must consider the system to
               be out of control.

12.1.3.5.2     If contamination is a problem, then the source of the
               contamination must be investigated and appropriate corrective
               measures must be taken and documented before further sample
               analysis proceeds.  It is the Contractor's responsibility to
               insure that method interferences caused by contaminants in
               solvents, reagents, glassware, and sample storage and sample
               processing hardware that lead to discrete artifacts and/or
               elevated baselines be investigated and appropriate corrective
               actions be taken and documented before further sample analysis.
               All samples associated with a contaminated sulfur cleanup blank
               must be re-extracted/reanalyzed at no additional cost to the
               Agency.

12.1.3.5.3     If the surrogate recoveries in the sulfur cleanup blank do not
               meet the acceptance criteria listed in Section 12.1.3.4.5,
               first reanalyze the sulfur cleanup blank.   If surrogate
               recoveries do not meet the acceptance criteria after
               reanalysis, the sulfur cleanup blank and all samples associated
               with that sulfur cleanup blank must be re-extracted and
               reanalyzed at no additional cost o the Agency.

12.1.3.5.4     If the sulfur cleanup blank failed to meet the criteria used in
               Sections 12.1.3.4.2 and 12.1.3.4.5, then there is an instrument
               problem.  Correct the instrument problem and reanalyze the
               sulfur cleanup blank.
                                   D-72/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 12
                                                               Quality Control
                                                                Blank Analyses
12.1.4
Instrument Blanks
12.1.4.1    Summary of Instrument Blanks

            An instrument blank is a volume of clean solvent spiked with the
            surrogates and' analyzed on each GC column and instrument used for
            sample analysis.  The purpose of the instrument blank is to
            determine the levels of contamination associated with the
            instrumental analysis itself, particularly with regard to the
            carry over of analytes from standards or highly contaminated
            samples into other analyses.

12.1.4.2    Frequency of Instrument Blanks

            The first analysis in a 12-hour analysis sequence must be an
            instrument blank.  All acceptable sample analyses are to be
            bracketed by acceptable instrument blanks, as described in Section
            10.2.2.1.  If more than 12 hours have elapsed since the injection
            of the instrument blank that>bracketed a previous 12-hour period,
            an instrument blank must be  analyzed to initiate a new 12-hour
            sequence.

12.1.4.3    Procedure for Instrument Blanks

12.1.4.3.1     Prepare the  instrument blank by spiking the surrogates into
               .hexane or iso-octane for  a concentration of 20 ng/mL of
               tetrachloro-m-xylene and  decachlorobiphenyl.

12.1.4.3.2     Analyze the  instrument blank according to Section 10.2 at the
               frequency listed in Section 12.1.4.2.

12.1.4.3.3     For comparing the results of the instrument blank analysis to
               the CRQLs, assume that the material in the  instrument resulted
               from  the extraction of a  1 L water sample and calculate the
               concentration of each analyte using the equation in Section
               11.2.1.1.  Compare the results to one-half  the GRQL values for
               water samples in Exhibit  C  (Pesticides-) .

12.1.4.4    Technical Acceptance Criteria for Instrument Blanks

12.1.4.4.1     The requirements below apply independently  to each GC Column
               and to all instruments used for  these analyses.  Quantitation
               must  be performed and reported independently  (on Form I PEST)
               for each GC  Column.

12.1.4.4.2     All instrument blanks must be prepared and  analyzed at the
               frequency described in Section 12.1.4.2, using the procedure in
               Section 12.1.4.3 on a GC/EC system meeting  the initial
               calibration  and calibration verification technical acceptance
               criteria.
12.1.4.4.3     The concentration of each of the target analytes  (Exhibit C
                (Pesticides)) in the instrument blank must  be less than 0.5
               times the CRQL for that  analyte.
                                    D-73/PEST
                                                               OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 12
Quality Control
Matrix Spike/Matrix Spike Duplicate  (MS/MSD)

12.1.4.4.4     The instrument blank must meet all sample technical acceptance
               criteria in Sections 11.3.4 to 11.3.7.

12.1.4.5    Corrective Action for Instrument Blanks

12.1.4.5.1     If analytes are detected at greater than half the CRQL, or the
               surrogate RTs are outside the RT windows, all data collection
               must be stopped, and corrective action must be taken.  Data for
               samples which were run between the last acceptable instrument
               blank and the unacceptable blank are considered suspect.  An
               acceptable instrument blank must be run before additional data
               are collected.  After an acceptable instrument blank is run,
               all samples which were considered.suspect as defined by the
               criteria described above must be reinjected during a valid run
               sequence at no additional cost to the Agency and must be
               reported.

12.2  Matrix Spike/Matrix Spike Duplicate  (MS/MSD)

12.2.1   Summary of MS/MSD

         In order to evaluate the effects of the sample matrix on the methods
         used for pesticide/Aroclor analyses, the Agency has prescribed a
         mixture of pesticide/Aroclor target compounds to be spiked into two
         aliquots of a sample, and analyzed in accordance with the appropriate
         method.

12.2.2   Frequency of MS/MSD Analysis

12.2.2.1    A matrix spike and matrix spike duplicate must-be extracted and
            analyzed for every 20 field samples of a similar matrix in an SDG.
            NOTE:  There is no differentiation between "low" and "medium"
            level soil/sediment samples in this method.  Therefore only one
            soil/sediment MS/MSD is to be submitted per Sample Delivery Group
            (SDG).  MS/MSD samples shall be analyzed unless otherwise
            specified on the Traffic Report  (TR).  If no MS/MSD samples are
            specified on the TR, the Contractor shall contact SMO to confirm
            that MS/MSD analyses are not required.

12.2.2.2    As part of the Agency's QA/QC program, water, rinsate samples
            and/or field blanks may be delivered to a laboratory for analysis.
            Do not perform MS/MSD analysis on a water rinsate sample or field
            blank.                                          =

12.2.2.3    If the EPA Region designates a sample to be used as an MS/MSD,
            then that sample must be used.  If there is insufficient sample
            .volume remaining to perform an MS/MSD, then the Contractor shall
            choose another sample to perform an MS/MSD analysis.  At the time
            the selection is made, the Contractor shall notify the Region
            (through SMO) that insufficient-sample was received and identify
            the EPA sample selected for the MS/MSD analysis.  The rationale
            for the choice of another sample other than the one designated by
            the Agency shall be documented in the SDG Narrative.
                                   D-74/PEST                            OLM04.2

-------
                                   Exhibit D Pestrcides/Aroclors -- Section 12
                                                               Quality Control
                                  Matrix Spike/Matrix Spike Duplicate (MS/MSD)

12.2.2.4    If there is insufficient sample volume remaining in any of the
            samples in an SDG to perform an MS/MSD, the Contractor shall
            immediately contact SMO to inform them of the problem.  SMO will
            contact the Region for instructions.   The Region will either
            approve that no MS/MSD be performed,  or require that a reduced
            sample aliquot be used for the MS/MSD analysis.  SMO will notify
            the Contractor of the Region's decision.  The Contractor shall
            document the decision in the SDG Narrative. .

12.2.2.5    If it appears that the Region has requested MS/MSD analysis at a
            greater frequency then required by the contract, the Contractor
            shall contact SMO.  SMO will contact the Region to determine which
            samples should have an MS/MSD performed on them.  SMO will notify
           . the Contractor of the Region's decision.  The Contractor shall
            document the decision in the SDG Narrative.  If this procedure is
            not followed, the Contractor will not be paid for MS/MSD analysis
            performed at a greater frequency than required by the contract.-

12.2.2.6    When a Contractor receives only performance evaluation  (PE)
            samples, no MS/MSD shall be performed within that SDG.

12.2.2.7    When a Contractor receives a PE sample as part of a larger SDG, a
            sample other than the PE sample must be chosen for the MS/MSD when
            the Region did not designate samples to be used for this purpose.
            If the PE-sample is received as an ampulated standard extract, the
            ampulated PE sample is not considered to be another matrix type.

12.2.3   Procedure for Preparing MS/MSD

12.2.3.1    Water Samples

            For water samples, measure out two additional  1 L aliquots of the
            sample chosen for spiking.  Adjust the pH of the samples  (if
            required) and fortify each with 1 mL of matrix spiking  solution.
            Using a syringe or volumetric pipet, add 1 mL  of surrogate spiking
            solution  to each  sample.  Extract, concentrate, cleanup, and
            analyze matrix spikes and matrix spike duplicate according to
            Section 10.0.

12.2.3.2    Soil/Sediment Samples

            For  soil/sediment samples weigh out two  additional  30  g (record
            weight to the nearest  0.1 g)  aliquots  of the sample  chosen for
            spiking.  Add 1 mL of matrix  spiking  solution  and 2 mL of
            surrogate  solution.  Extract, concentrate, cleanup,  and analyze
            matrix spikes and matrix spike duplicates  according  to Section
            10.0.

12.2.3.3    Note:  Before any MS/MSD analysis, analyze the original sample,
            then analyze  the  MS/MSD  at  the  same concentration as  the most
            concentrated  extract  for which  the original  sample  results will be
            reported.   For example,  if  the original  sample is to  be reported
            at a 1:1  dilution and  a  1:10  dilution,  then  analyze  and report  the
            MS/MSD at a 1:1  dilution only.  However, if  the original sample  is
            to be reported at a  1:10 dilution  and a 1:100  dilution,  then  the


                                   D-75/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 12
Quality Control
Matrix Spike/Matrix Spike Duplicate  (MS/MSD)

            MS/MSD must be analyzed and reported at a 1:10 dilution only.  Do
            not further dilute the MS/MSD samples to get either spiked or
            nonspiked analytes within calibration range.

12.2.4   Calculations for MS/MSD

12.2.4.1    The percent recoveries and the relative percent difference between
            the recoveries of each of the compounds in the matrix spike
            samples will be calculated and reported by using the following
            equations:

            EQ. 18

                                            CCTD — GT?
                     Matrix Spike Recovery = 	 x 100
                                               SA-
            Where,

            SSR
            SR
            SA
Spike' sample result
Sample result
Spike added
            EQ. 19
            Where,

            RPD
            MSR
            MSDR  =
                           RPD =
             [MSR - MSDR|
           -  (MSR + MSDR)
                                               x 100
Relative percent difference
Matrix spike recovery
Matrix spike duplicate recovery
12.2.4.2    The vertical bars in the formula above indicate the absolute value
            of the difference, hence RPD is always expressed as a positive
            value.

12.2.5   Technical Acceptance Criteria for MS/MSD

12.2.5.1    The requirements below apply independently to each GC column and
            to all instruments used for these analyses.  Quantitation must be
            performed on both GC columns.

12.2.5.2    All MS/MSD must be prepared and analyzed at the frequency
            described in Section 12.2.2, using the procedure above and in
            Section 10 on a GC/EC system meeting the initial calibration,
            calibration verification, and blank technical acceptance criteria.
            MS/MSD must be cleaned-up, when required, on a GPC meeting the
            technical acceptance criteria for GPC calibration and GPC
            calibration checks.  MS/MSD must be cleaned-up using florisil
            meeting the technical acceptance criteria for florisil.  MS/MSD
            must be bracketed at 12-hour intervals (or less) by acceptable
                                   D-76/PEST
                                                   OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors --  Section 12
                                                               Quality Control
                                  Matrix Spike/Matrix Spike Duplicate (MS/MSD)

            analyses of instrument blanks, PEMS,  and individual standard
            mixtures A and B as described in Section 10.2.2.1.

12.2.5.3    The samples must be extracted and analyzed within the contract
            required holding times.

12.2.5.4    The retention time for each of the surrogates must be within the
            retention time window as calculated in Section 9 for both GC
            columns.

12.2.5.5    The limits for matrix spike compound recovery and RPD are given in
            Table 3.  As these limits are only advisory, no further action by
            the laboratory is required.  However, frequent failures to meet
            the limits for recovery or RPD warrant investigation by the
            laboratory, and may result in questions from the Agency.

12.2.6   Corrective Action for MS/MSD

         Any MS/MSD which fails to meet the technical acceptance criteria for
         MS/MSD must be reanalyzed at no additional cost to the Agency.
                                   D-77/PEST                            OLM04.2

-------
Exhibit D Pesticicies/Arociors -- Sections 13-16
Method Performance/Pollution Prevention/Waste Management/References

13.0  METHOD PERFORMANCE

      Not Applicable

14.0  POLLUTION PREVENTION

14.1  Pollution prevention encompasses any technique that reduces or
      eliminates the quantity or toxicity of waste at the point of generation.
      Numerous opportunities for pollution prevention exist in laboratory
      operation.  The EPA has established a preferred hierarchy of
      environmental management techniques that places pollution prevention as
      the management option of first choice.  Whenever feasible, laboratory-
      personnel should use pollution prevention techniques to address their
      waste generation.  When wastes cannot be feasibly reduced at the source,
      the Agency recommends recycling as the next best option.

14.2  For information about pollution prevention that may be applicable to
      laboratories and research institutions, consult Less is Better:
      Laboratory Chemical Management for Waste Reduction, available from the
      American Chemical Society's Department-of Government Relations and
      Science Policy, 1155 16th Street N.W., Washington, D.C.  20036,
       (202)872-4477.

15.0  WASTE MANAGEMENT
                                  t
      The Environmental Protection Agency requires that laboratory waste
      management practices be conducted consistent with all applicable rules
      and regulations.  The Agency urges laboratories to protect the air,
      water, and land by minimizing and controlling all releases from hoods
      and bench operations, complying with the letter and spirit of any sewer
      discharge permits and regulations, and by complying with all solid and
      hazardous waste regulations, particularly the hazardous waste
      identification rules and land disposal restrictions.  For further
      information on waste management consult The Waste management Manual for
      Laboratory Personnel, available from the American Chemical Society at
      the address listed in Section 14.2.  .

16.0  REFERENCES

      Not Applicable
                                   D-78/PEST                            OLM04.2

-------
                                   Exhibit D Pesticides/Aroclors -- Section 17
                                                    Tables/Diagrams/Flowcharts
17.0  TABLES/DIAGRAMS/FLOWCHARTS
                                    Table  1
                    Retention Time Windows for Single and
                    Multicomponent Analytes and Surrogates
Compound
alpha-BHC
beta-BHC
gamma -BHC (Lindane)
delta-BHC
Heptachlor
Aldrin }
alpha - Chlordane
gamma- Chlordane
Heptachlor epoxide
Dieldrin
Endrin
Endrin aldehyde
Endrin ketone
4,4' -ODD
4,4' -DDE
4,4 '-DDT
Endosulfan I
Endosulfan II
Endosulfan sulfate
Methoxychlor
Aroclors
Toxaphene
Tetrachloro-m-xylene
Decachlorobiphenyl
Retention Time Window (minutes)
± 0.05
±0.05
±0.05
±0.05
± 0.05
±0.05
±0.07
±0.07
±0.07
±0.07
± 0.07
±0.07
± 0.07
±0.07
± 0.07
±0.07
±0.07
± 0.07
±0.07
± 0.07
± 0.07
..± 0.07
±0.05
± 0.10
                                    D-79/PEST                            OLM04.2

-------
Exhibit D Pesticides/Aroclors -- Section 17
Tables/Diagrams/Flowcharts
                                    Table  2
                    Number of Potential Quantitation Peaks
 Multicomponent Analyte
 No. of Potential

Quantitation Peaks
 Aroclor 1016/1260

 Aroclor 1221

 Aroclor 1232

 Aroclor 1242

 Aroclor 1248

 Aroclor 1254

 Toxaphene
        5/5

         3

         4

         5

         5

         5

         4
                                    Table 3
                           Matrix Spike Recovery and
                      Relative Percent Difference Limits
Compound
gamma -BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4, 4 --DDT
%Recovery
Water
56-123
40-131
40-120
52-126
56-121
38-127
RPD
Water
15
20
22
18
21
27
%Recovery
Soil
46-127
35-130
34-132
31-134
42-139
23-134
RPD
Soil
50
31
43
38
45
50
                                   D-80/PEST
                            OLM04.2

-------
                          EXHIBIT E
QUALITY ASSURANCE/QUALITY CONTROL PROCEDURES AND REQUIREMENTS
                             E-l                               OLM04.2

-------
  Exhibit E  - Quality Assurance/Quality Control Procedures and Requirements

                              Table of Contents


Section                                                                   Page

1.0   OVERVIEW	4

2.0   INTRODUCTION	  5

3.0   QUALITY ASSURANCE PLAN	7
      3.1   Introduction	7
      3.2   Required Elements of a Quality Assurance Plan	  7
      3.3   Updating and Submitting the Quality Assurance Plan	9
      3.4   Corrective Actions  	 10

4.0   STANDARD OPERATING PROCEDURES 	 11
      4.1   Introduction	11
      4.2   Format	12
      4.3   Requirements	12
      4.4   Submitting and Updating SOPs	15
      4.5   Corrective Actions  	 17

5.0   ANALYTICAL STANDARDS REQUIREMENTS 	 18
      5.1   Overview	;	18
      5.2   Preparation of Chemical Standards from the Neat High Purity
            Bulk Material	18
      5.3   Purchase  of Chemical Standards Already in Solution  	 19
      5.4   Requesting Standards From the EPA Standards Repository   .... 22
      5.5   Documentation  of the Verification and Preparation of Chemical
            Standards	22
      5.6   Corrective Actions	23

6.0   DETERMINATION OF METHOD EQUIVALENCY FOR ALTERNATIVE EXTRACTION
      PROCEDURES	24
      6.1   Initial Precision Recovery  (IPR) Study	-  ... 24
      6.2   Analytical Protocol Required   	  	 25
      6.3   Quantit'ation Limits/Quality Control Requirements	25
      6.4   Data Deliverable Requirements  	 26

7.0   CONTRACT COMPLIANCE  SCREENING  	 27

8.0   REGIONAL DATA REVIEW	28

9.0   PROFICIENCY  TESTING	29
      9.1    Performance Evaluation Samples	 29
      9.2   Quarterly Blind Audits   .  . •	30

10.0  GC/MS AND GC/EC TAPE AUDITS  ;  .  .	33
      10.1  Overview	33
      10.2   Submission of  the GC/MS and GC/EC  Tapes	35
      10.3   Responding to  the GC/MS and GC/EC  Tape Audit Report	35
      10.4   Corrective Actions   	 36


                                      E-2                              OLM04.2

-------
11.0  DATA PACKAGE AUDITS	38
      11.1  Overview	38
      11.2  Responding to the Data Package Audit Report	38
      11.3  Corrective Actions	  38

12.0  ON-SITE LABORATORY EVALUATIONS  	  39
      12.1 • Overview	-39
      12.2  Quality Assurance On-Site Evaluation  .	  39
      12.3  Evidentiary Audit •	39
      12.4  Discussion of the On-Site Team's Findings 	  40
      12.5  Corrective Action Reports for Follow-Through to Quality
            Assurance and Evidentiary Audit Repprts  	  ...  40
      12.&  Corrective Actions  	  41

13.0  QUALITY ASSURANCE AND DATA TREND ANALYSIS .	  42

14.0  DATA MANAGEMENT	43
                                      E-3  .                             OLM04.2

-------
Exhibit E -- Section 1
Overview
1.0   OVERVIEW

1.1   Quality assurance and quality control are integral parts of the
      Environmental Protection Agency's  (EPA) Contract Laboratory Program
      (CLP).  The quality assurance (QA) process consists of management review
      and oversight at the planning, implementation, and completion stages of
      the environmental data collection activity, and ensures that data
      provided are of the quality required.  The quality control (QC) process
      includes those activities required during data collection to produce the
      data quality desired and to document the quality of the collected data.

1.2   During the planning of an environmental data collection program, QA
      activities focus on defining data quality criteria and designing a QC
      system to measure the quality of data being generated.  During the
      implementation of the data collection effort, QA activities ensure that
      the QC system is functioning effectively, and that the deficiencies
      uncovered by the QC system are corrected.  After environmental data are
      collected, QA activities focus on assessing the quality of data obtained
      to determine its suitability to support enforcement or remedial
      decisions.

1.3   This exhibit describes the overall quality assurance/quality control
      operations and the processes by which the CLP meets the QA/QC objectives
      defined above.  This contract requires.a variety of QA/QC activities.
      These contract requirements are the minimum QC operations necessary to
      satisfy the analytical requirements associated with the determination of
      the different method analytes.  These QC operations are designed to
      facilitate laboratory comparison by providing EPA with comparable data
      from all Contractors.  These requirements do not release the analytical
      Contractor from maintaining their  own QC checks on method and instrument
      performance.
                                      E-4                               OLM04.2

-------
                                                        Exhibit E -- Section 2
                                                                  Introduction
2.0   INTRODUCTION

2.1   Appropriate use of data generated under the large range of analytical
      conditions encountered in environmental analyses requires reliance on
      the QC procedures and criteria incorporated into the methods.  The
      methods in this contract have been validated on samples typical of those
      received by the laboratories in the CLP.  However, the validation of
      these methods does not guarantee that they perform equally well for all
      sample matrices encountered.  Inaccuracies can also result from causes
      other than unanticipated matrix effects, such as sampling artifacts,
      equipment malfunctions, and operator error.  Therefore, the quality
      control component of each method is indispensable.

2.2   The data acquired from QC procedures are used to estimate and evaluate
      the information content of analytical results and to determine the
      necessity for or the effect of corrective action procedures.  The
      parameters used to estimate information content include precision,
      accuracy, detection limit, and other quantitative and qualitative
      indicators.  In addition, QC procedures give an overview of the
      activities required in an integrated program to generate data of known
      and documented quality required to meet defined objectives.

2.3   The necessary components of a complete QA/QC program include internal QC
      criteria that demonstrate acceptable levels of performance, as
      determined by QA review.  External review of data and procedures is
      accomplished by the monitoring activities of the National Program
      Office, Regional data users, Sample Management Office  (SMO), and the
      Quality Assurance Technical Support  (QATS) Laboratory.  Each external
      review accomplishes a different purpose.  These reviews are described in
      specific sections of this exhibit.  Laboratory evaluation samples, GC/MS
      and GC/EC tape audits, and data packages provide an external QA
      reference for the program.  A Contractor on-site evaluation system is
      also part of the external QA monitoring.  A feedback loop provides the
      results of the various review functions to the Contractors through
      direct communications with the Technical Project Officers  (TPOs) and
      Administrative Project Officers (APOs).

2.4   This exhibit does not provide specific instructions for constructing QA
      plans, QC systems, or a QA organization.  It is, however, an explanation
      of the QA/QC requirements of the program.  It outlines some minimum
      standards for QA/QC programs.  It also includes specific items that are
      required in a QA plan and by the QA/QC documentation detailed in this
      contract.  Delivery o'f this documentation provides the Agency with a
      complete data package which will stand alone, and limits the need for
      contact with the Contractor or with an analyst, at a later date, if some
      aspect of the analysis is questioned.

2.5   In order to assure that the product delivered by the Contractor meets
      the requirements of the contract,  and to improve interlaboratory data
      comparison, the Agency requires the following from the Contractor:

      •   Preparation of and adherence to a written quality assurance plan,
          the elements of which are designated in Section 3,
                                      E-5                              OLM04.2

-------
Exhibit E -- Section 2
Introduction
      •   Preparation of and adherence to QA/QC standard operating procedures
          as described in Section 4,

      •   Adherence to the analytical methods and associated QC requirements
          specified in the contract,

      •   Verification of analytical standards and documentation of the purity
          of neat materials and the purity and accuracy of solutions obtained
          from private chemical supply houses,

      •   Submission of all raw data and pertinent documentation for Regional
          review,

      •   Participation in the analysis of laboratory evaluation samples,
          including adherence to corrective action procedures,

      •   Submission, upon request, of GC/MS and/or GC/EC tapes and applicable
          documentation for tape audits, including a copy of the sample data
          package,

      •   Participation in on-site  laboratory evaluations, including adherence
          to corrective action procedures, and

      •   Submission of all original documentation generated during sample
          analyses for Agency review.
                                      E-6                               OLM04.2

-------
                                                        Exhibit E -- Section 3
                                                        Quality Assurance Plan
3.0   QUALITY ASSURANCE PLAN

3.1   Introduction.  The Contractor shall establish a quality assurance
      program with the objective of providing sound analytical chemical
      measurements.  This program shall incorporate the quality control
      procedures, any necessary corrective action, and all documentation
      required during data collection as well as the quality assessment
      measures performed by management to ensure acceptable data production.

3.1.1    As evidence of such a program, the Contractor shall prepare a written
         quality assurance plan  (QAP) which describes the procedures that are
         implemented to achieve the following:

         •   Maintain data integrity, validity, and usability,

         •   Ensure that analytical measurement systems are maintained in an
             acceptable state of stability and reproducibility,

         •   Detect problems through data assessment and establish corrective
             action procedures which keep the analytical process reliable, and

         •   Document all aspects of the measurement process in order to
             provide data which are technically sound and legally defensible.

3.1.2    The QAP shall present, in specific terms, the policies, organization,
         objectives, functional guidelines, and specific QA and QC activities
         designed to achieve the data quality requirements in this contract.
         Where applicable, standard operating procedures pertaining to each
         element shall be included or referenced as part of the QAP.  The QAP
         shall be paginated consecutively in ascending order.  The QAP shall
         be available during on-site laboratory evaluations.  Additional
         information relevant to the preparation of a QAP can be found in
         Agency and American Society for Testing and Materials publications.

3.2   Required Elements of a Quality Assurance Plan.  The required elements of
      a laboratory's QAP are outlined in this section.  This outline should be
      used as a framework for developing the QAP.

      A.  Organization and Personnel

          1.  QA Policy and Objectives

          2.  QA Management

              a.  Organization

              b.  Assignment of QC and QA Responsibilities

              c.  Reporting Relationships

              d.  QA Document Control Procedures

              e.  QA Program Assessment Procedures


                                      E-7                               OLM04.2

-------
Exhibit E -- Section 3
Quality Assurance Plan
          3.  Personnel

              a.  Resumes

              b.  Education and Experience

              c.  Training Progress

      B.  Facilities and Equipment

          1.  Instrumentation and Backup Alternatives

          2.  Maintenance Activities and Schedules

      C.  Document Control

          1.  Contractor Notebook Policy

          2.  Sample Tracking/Custody  Procedures

          3.  Logbook  Maintenance and  Archiving Procedures'

          4.  Case File Organization,  Preparation,  and  Review  Procedures

          5.  Procedures for Preparation, Approval,  Review,  Revision,  and
              Distribution of Standard Operating  Procedures

          6.  Process  for Revision  of  Technical or  Documentation Procedures

      D.  Analytical Methodology

          1.  Calibration Procedures and  Frequency

          2.  Sample Preparation/Extraction Procedures

          3.  Sample Analysis Procedures

          4.  Standards  Preparation Procedures

          5.  Decision Processes,  Procedures,  and Responsibility for
               Initiation of  Corrective Action

       E.  Data  Generation

           1.  Data Collection Procedures

          2.  Data Reduction Procedures

           3.   Data Validation  Procedures

           4.   Data Reporting and Authorization Procedures




                                       E-8                               OLM04.2

-------
                                                          Exhibit E -- Section 3
                                                          Quality Assurance Plan


        F.   Quality Control

            1.   Solvent, Reagent, and Adsorbent Check Analysis

            2.   Reference Material Analysis

            3.   Internal Quality Control Checks

            4.   Corrective Action and Determination of QC Limit Procedures

            5.   Responsibility Designation

        G.   Quality Assurance

            1.   Data Quality Assurance

            2.   Systems/Internal Audits

            3.   Performance/External Audits

            4.   Corrective Action Procedures

            5.   Quality Assurance Reporting Procedures

            6.   Responsibility Designation

  3.3   Updating and Submitting the Quality Assurance Plan

  3.3.1    Initial Submission.  During the contract solicitation process, the
|           Contractor is required to submit their QAP to the Contracting
           Officer.  Within 60 days after contract award, the Contractor shall
           revise the QAP to be in full compliance with the requirements of this
           contract.  The Contractor shall maintain the QAP on file at the
           Contractor's facility for the term of the contract.  The revised QAP
           will become the official QAP under the contract and may be used
           during legal proceedings.  Both the initial QAP submission and the
           revised QAP shall be paginated consecutively in ascending order.  The
           revised QAP shall include:

           •   Changes resulting from  (1) the Contractor's internal review of
               their organization, personnel, facility, equipment, policy, and
               procedures and  (2) the Contractor's implementation of the
               requirements of the contract, and

           •   Changes resulting from the Agency's review of the laboratory
               evaluation sample data, bidder-supplied documentation, and
               recommendations made during the pre-award on-site laboratory
               evaluation.

|  3.3.1.1     The Contractor shall send a copy of the latest version of the QAP
              within 7 days of a"request from a Technical Project Officer or
              Administrative Project Officer.  The Agency requestor will
              designate the recipients.


                                        E-9                               OLM04.2

-------
Exhibit E -- Section 3
Quality Assurance Plan
3.3.2    Subsequent Updates and Submissions.  During the term of the contract,
         the Contractor shall amend the QAP when the following circumstances
         occur:

         •   The Agency modifies the contract,

         •   The Agency notifies the Contractor of deficiencies in the QAP,

         •   The Agency notifies the Contractor of deficiencies resulting  from
             the Agency's review of the Contractor's performance,

         •   The Contractor identifies deficiencies resulting from the
             internal review of the QAP,

         •   The Contractor's organization, personnel,  facility, equipment,
             policy, or procedures change, or

         •   The Contractor identifies deficiencies resulting from the
             internal review of changes in their organization, personnel,
             facility, equipment, policy, or procedures.

3.3.2.1      The Contractor shall amend the QAP within  30 days of when the
             circumstances listed above result in a discrepancy between what
             was previously described  in the QAP and what is presently
             occurring at the Contractor's facility.  When  the QAP is amended,
             all changes in the QAP shall be clearly marked (e.g., a bar  in the
             margin indicating where the change is found in the document, or
             highlighting the change by underlining the change, bold printing
             the change, or using a different print font)..  The amended pages
             shall  have the date on which the changes were  implemented.   The
             Contractor shall incorporate all amendments to the latest version
             of the QAP.  The Contractor shall archive  all  amendments to  the
             QAP for future reference  by the Agency.

3.3.2.2      The Contractor shall send a copy of the  latest version  of  the  QAP
             within 7 days of a request  from a Technical Project  Officer  or
             Administrative Project Officer.  The Agency requestor will
             designate  the recipients.

3.4   Corrective Actions.  If the  Contractor  fails to  adhere  to  the
       requirements listed  in Section  3, the Contractor may expect, but  the
       Agency is not limited  to,  the  following actions: reduction of numbers
       of samples  sent  under  this  contract,  suspension  of sample  shipment to
       the Contractor,  a GC/MS and/or GC/EC tape audit, a data package audit,
       an on-site  laboratory  evaluation, a  remedial laboratory evaluation
       sample,  and/or contract sanctions.
                                      E-10                              OLM04.2

-------
                                                         Exhibit E --  Section 4
                                                  Standard Operating Procedures
 4.0    STANDARD OPERATING PROCEDURES

 4.1    Introduction.   In order to obtain reliable results,  adherence to
       prescribed analytical methodology is imperative.   In any operation that
       is performed on a repetitive basis,  reproducibility is best accomplished
       through the use of standard operating procedures  (SOPs).   As defined by
       EPA,  an SOP is a written document which provides  directions for the
       step-by-step execution of an operation,  analysis,  or action which is
       commonly accepted as the method for performing certain routine or
       repetitive tasks.

 4.1.1    SOPs prepared by the Contractor shall be functional (i.e., clear,
          comprehensive, up-to-date, and sufficiently detailed to permit
          duplication of results by qualified analysts).   The SOPs shall be
          paginated consecutively in ascending order.

 4.1.2    All SOPs shall reflect activities as they are  currently performed by
          the Contractor.  In addition,  all SOPs shall be:

          •    Consistent with current Agency regulations,  guidelines, and the
              CLP contract's requirements.

          •    Consistent with instrument manufacturers'  specific instruction
              manuals.

          •    Available to the Agency during an on-site  laboratory evaluation.
              A complete set of SOPs shall be bound together and available for
              inspection at such evaluations.  During on-site evaluations,
              Contractor personnel may be asked to demonstrate the application .
              of the SOPs.

          •    Available to the designated recipients within 7 days, upon
              request by the Technical Project Officer or Administrative
              Project Officer.

          •    Capable of providing for the development of documentation that is
              sufficiently complete to record the performance of all tasks
              required by the.protocol.

          •    Capable of demonstrating the validity of data reported by the
              Contractor and explaining the cause of missing or inconsistent
              results.

          •    Capable of describing the corrective measures and feedback
              mechanism utilized when analytical results do not meet protocol
              requirements.

          •    Reviewed regularly and updated as necessary when contract,
              facility, or Contractor procedural modifications are made.

          •    Archived for future reference in usability or evidentiary
              situations.

          •    Available at specific work stations as appropriate.

:      '               .                . E-ll                              OLM04.2

-------
Exhibit E -- Section 4
Standard Operating Procedures
         •   Subject to a document control procedure which precludes the use
             of outdated or inappropriate SOPs.

4.2   Format.  The format for SOPs may vary depending upon the kind of
      activity for which they are prepared; however, at a minimum, the
     vfollowing sections shall be included:

      •   Title page,

      •   Scope and application,

      •   Definitions,

      •   Procedures,

      •   QC limits,

      •   Corrective action procedures, including procedures for  secondary
          review of information being generated,

      •   Documentation description and example forms,

      •   Miscellaneous notes and precautions, and

      •   References.

4.3   Requirements.  The Contractor shall maintain  the  following  SOPs.

4.3.1     Evidentiary SOPs  for required chain-of-custody and document control
          are discussed in  Exhibit F.

4.3.2     Sample Receipt and Storage

          •   Sample receipt and identification  logbooks

          •   Refrigerator  temperature logbooks

          •   Extract storage  logbooks

          •   Security  precautions

4.3.3     Sample Preparation

          •   Reagent purity check procedures and  documentation

          •   Extraction procedures

          •   Extraction bench sheets

          •   Extraction logbook maintenance

4.3.4     Glassware Cleaning

4.3.5     Calibration  (Balances)

                                      E-12               '                OLM04.2

-------
                                                        Exhibit E -- Section 4
                                                 Standard Operating Procedures
         •   Procedures

         •   Frequency requirements

         •   Preventative maintenance schedule and procedures

         •   Acceptance criteria 'and corrective actions

         •   Logbook maintenance

4.3.6    Analytical Procedures  (for each Analytical System, including GPC)

         •   Instrument performance specifications

         •   Instrumental operating procedures

         •   Data acquisition system operation

         •   Procedures when automatic quantitation algorithms are overridden

         •   QC required parameters

         •   Analytical run/injection logbooks

         •   Instrumental error and editing flag descriptions and resulting
             corrective actions

4.3.7    Maintenance Activities  (for each Analytical System, including GPC)

         •   Preventative maintenance schedule and procedures

         •   Corrective maintenance determinants and procedures

         •   Maintenance authorization

4.3.8    Analytical Standards

         •   Standard coding/identification and inventory system

         •   Standards preparation logbook(s)

         •   Standards preparation procedures

         •   Procedures for equivalency/traceability analyses and
             documentation

         •   Purity logbook  (primary standards and solvents)

         •   Storage, replacement, and labeling requirements

         •   QC and corrective action measures



                                    • E-13                              OLM04.2

-------
Exhibit E -- Section 4
Standard Operating Procedures
4.3.9    Data Reduction Procedures

         •   Data processing systems operation

         •   Outlier identification methods

         •   Identification of data requiring corrective action

         •   Procedures for format and/or forms for each operation

4.3.10   Documentation Policy/Procedures

         •   Contractor/analysts' notebook policy, including review policy

         •   Complete SDG File contents

         •   Complete SDG File organization and assembly procedures,  including
             review policy

         •   Document inventory procedures, including review policy

4.3.11   Data Validation/Self-Inspection Procedures

         •   Data  flow and chain-of-command for data review

         •   Procedures for measuring precision and accuracy

         •   Evaluation parameters  for  identifying systematic  errors

         •   Procedures to ensure that  hardcopy and electronic deliverables
              (e.g., diskette,  telefacsimile)  are  complete  and  compliant  with
             the requirements  in Exhibits B and'H

         •   Procedures to ensure that  hardcopy deliverables are in agreement
             with  their comparable  electronic deliverables

         •   Demonstration of  internal  QA inspection procedure (demonstrated
             by supervisory  sign-off on personal  notebooks,  internal
             performance  evaluation samples,  etc.)

          •   Frequency and type of  internal audits  (e.g.,  random,  quarterly,
             spot  checks, perceived trouble areas)

          •   Demonstration of  problem  identification,  corrective actions and
             resumption of analytical  processing; sequence resulting  from
             internal audit  (i.e.,  QA  feedback)

          •   Documentation of  audit reports  (internal  and external),  audit
             response,  corrective  action, etc.
                                      E-14                              OLM04.2

-------
                                                          Exhibit E -- Section 4
                                                   Standard Operating Procedures


  4.3.12   Data Management and Handling

           •    Procedures for controlling and estimating data entry errors

           •    Procedures for reviewing changes to data and deliverables and
               ensuring traceability of updates

           •    Life cycle management procedures for testing, modifying and
               implementing changes- to existing computing systems including
               hardware, software, and documentation or installing new systems

           •    Database security, backup and archival procedures including
               recovery from system failures

           •    System maintenance procedures and response time

           •    Individuals(s) responsible for system operation, maintenance,
               • data integrity and security

           •    Specifications for staff training procedures

           •    Storage, retrieval, and verification of the completeness and
               readability of GC/MS and GC/EC files transferred to magnetic
               media

  4;.. 4   Submitting and Updating SOPs

  4 .-;4.1    Initial Submission.  During the contract solicitation process, .the
|           Contractor is required to submit their SOPs to the Contracting
 •-,•:-•       Officer.  Within 60 days after contract award, the Contractor shall
   •j.       prepare and maintain on file, at their facility, a complete, revised
           set of SOPs fully compliant with the requirements of this contract.
           The revised SOPs will become the official SOPs under the contract and
           may be used during legal proceedings.  Both the initial submission of
           SOPs and the revised SOPs shall be paginated consecutively in
           ascending order.  The revised SOPs shall include:

           •    Changes resulting from  (1) the Contractor's internal review of
               their procedures and  (2) the Contractor's implementation of the
               requirements of the contract, and

           •    Changes resulting from the Agency's review of the laboratory
               evaluation sample data, bidder-supplied documentation, and
               recommendations made during the pre-award on-site laboratory
               evaluation.

|  4.4.1.1     The Contractor shall send a complete set of the latest version of
              SOPs or individually requested SOPs within 7 days of a request
              from a Technical Project Officer or Administrative Project
              Officer.  The Agency requestor will designate the recipients.

  4.4.2    Subsequent Updates and Submissions.  During the term of the contract,
           the Contractor shall amend the SOPs when the following circumstances
           occur:


                                       E-15                              OLM04.2

-------
 Exhibit E  --  Section  4
 Standard Operating  Procedures
           •    The  Agency modifies  the  contract,

           •    The  Agency notifies  the  Contractor of  deficiencies in their SOPs,

           •    The  Agency notifies  the  Contractor of  deficiencies resulting from
               the  Agency's review  of the Contractor's  performance,

           •    The  Contractor's procedures change,

           •    The  Contractor identifies deficiencies resulting from the
               internal review of their SOPs documentation,  or

           •    The  Contractor identifies deficiencies resulting from the
               internal review of their procedures.

  4.4.2.1     Existing SOPs shall be amended or new SOPs shall be written within
              30 days of when the circumstances listed above result in a
              discrepancy between what  was previously described in the SOPs and
              what  is presently occurring at the Contractor's facility.  All
              changes in the SOPs shall be clearly marked (e.g., a bar in the
              margin indicating where the change is in the document, or
              highlighting the change by underlining the change, bold printing
              the change, or using  a different print font).   The amended/new
              SOPs  shall have the date  on which the changes were implemented.

  4.4.2.2     When existing SOPs are amended or new SOPs are written, the
              Contractor shall document the reason(s) for the change, and
              maintain the amended or new SOPs -on f-ile at the laboratory
              facility.  Documentation of the reason(s) for the changes shall be
              maintained on file with the amended SOPs or new SOPs.

|  4.4.2.3     The Contractor shall  send a complete set of ttie latest version of
              SOPs or individually requested SOPs within 7 days of a request
              from a Technical Project Officer or Administrative Project
              Officer.  The Agency requestor will designate the recipients.    ,

  4.4.2.4     Documentation of the reason(s) for changes to the SOPs shall also
              be submitted with the SOPs.  An alternate delivery schedule for
              submitting the amended/new SOPs and their documentation  may be
              proposed by the Contractor, but it is the sole decision of the
              Agency, represented either by the Technical Project Officer or
              Administrative Project Officer, to approve or disapprove the
              alternate delivery schedule.  If an alternate delivery schedule is
              proposed, the Contractor shall describe in a letter to the
              Technical Project Officer, Administrative Project Officer, and the
              Contracting Officer why he/she is unable to meet  the delivery
              schedule listed in this section.  The Technical Project
              Officer/Administrative Project Officer will not grant an extension
              for greater than 30 days for amending/writing new SOPs.  The
              Technical Project Officer/Administrative Project  Officer will not
              grant  an extension for greater than 14 days for submission of the
              letter documenting the reasons for the changes and  for submitting
              amended/new SOPs.  The Contractor shall proceed and not assume

                                       E-16                              OLM04.2

-------
                                                        Exhibit E -- Section 4
                                                 Standard Operating Procedures
            that an extension will be granted until so notified by the
            Technical Project Officer and/or Administrative Project Officer.

4.5   Corrective Actions.  If the Contractor fails to adhere to the
      requirements listed in Section 4, the Contractor may expect, but the
      Agency is not limited to,  the following action:  reduction of number of
      samples sent under this contract, suspension of sample shipment to the
      Contractor, a GC/MS and/or GC/EC tape audit, a data package audit, an
      on-site laboratory evaluation, a remedial laboratory evaluation sample,
      and/or contract sanctions.
                                     E-17                              OLM04.2

-------
Exhibit E -- Section 5
Analytical Standards Requirements
5.0   ANALYTICAL STANDARDS REQUIREMENTS

5.1   Overview.  EPA will not supply analytical reference standards either for
      direct analytical measurements or for the purpose of traceability.  All
      Contractors shall be required to prepare from neat materials or purchase
      from private chemical supply houses those standards necessary to
      successfully and accurately perform the analyses required in this
      protocol.

5.2   Preparation of Chemical Standards from the Neat High Purity Bulk
      Material.  A Contractor may prepare their chemical standards from neat
      materials.  Contractors shall obtain the highest purity possible when
      purchasing neat chemical standards; when standards are purchased at less
      than 97% purity, the Contractor shall document the reason why a higher
      purity could not be obtained.

5.2.1    Neat  chemical standards shall be kept refrigerated when not being
         used  in the preparation of standard solutions.  Proper storage of
         neat  chemicals is essential in order to safeguard them from
         decomposition.

5.2.2    The purity of a compound can sometimes be misrepresented by a
         chemical supply house.  Since knowledge of purity is needed to
         calculate the concentration of solute in a solution standard,  it is
         the Contractor's responsibility to have analytical documentation
         ascertaining that the purity of each compound is correctly stated.
         Purity confirmation, when performed, should use either differential
         scanning calorimetry, gas chromatography with flame ionization
         detection, high performance liquid chromatography, infrared
         spectrometry, or other appropriate techniques.  Use of two or  more
         independent methods  is recommended.  The correction factor for
         impurity when weighing neat materials in the preparation of solution
         standards is:
          EQ.  1
                 .  , .   , .              ,   weight of pure compound
               weight of impure compound = —T-=	—	——,—
                                            [percent purity/100)
          where "weight of pure compound"  is that  required to prepare  a
          specific volume of a standard solution at  a specified concentration.

 5.2.3     When compound purity is assayed  to be 97%  or greater,  the  weight  may
          be used without correction to calculate  the concentration  of the
          stock standard.  If the compound purity  is assayed to be less than
          97%,  the weight shall be corrected when  calculating the concentration
          of the stock solution.

 5.2.4     Mis-identification of compounds  occasionally occurs and it is
          possible that a mislabeled compound may  be received from a chemical
          supply house.  It is the Contractor's responsibility to have
          analytical documentation ascertaining that all compounds used in  the
          preparation of solution standards are correctly identified.

                                      E-18                         -     OLM04.2

-------
                                                        Exhibit E -- Section 5
                                             Analytical Standards Requirements


         Identification confirmation, when performed, shall use gas
         chromatography/mass spectrometry analysis on at least two different
         analytical columns, or other appropriate techniques.

5.2.5    Calculate the weight of material to be weighed out for a specified
         volume taking into account the purity of the compound and the desired
         concentration.  A second person shall verify the accuracy of the
         calculations.  Check balances for accuracy with a set of standard
         weights every 12 hours.  All weighing shall be performed on an
         analytical balance to the nearest 0.1 mg and verified by a second
         person.  The solvent used to dissolve the solute shall be compatible
         with the protocol in which the standard is to be used; the solute
         shall be soluble, stable, and nbnreactive with the solvent.  In the
         case of a multicomponent solution, the components must not react with
         each other.

5.2.6    Transfer the solute to a volumetric flask and dilute to the specified
         solution volume with solvent after ensuring dissolution of the solute
         in the solvent.   Sonication or warming may be performed to promote
         dissolution of the solute.  This solution shall be called the primary
         standard and all subsequent dilutions shall be traceable back to the
         primary standard.

5.2.7    Log notebooks shall be kept for all weighing and dilutions.  All
     -•.V   subsequent dilutions from the primary standard and the calculations
     -•'.   for determining their concentrations shall be recorded and verified
         by a second person.  All solution standards shall be refrigerated
     .;,•   when not in use.  All solution standards shall be clearly labeled as
     ;f   to the identity of the compound or compounds, concentration, date
     :'•'•   prepared, solvent, and initials of the preparer.

5.3   Purchase of Chemical Standards Already in Solution.  Solutions of
      analytical reference standards can be purchased by"Contractors provided
      they meet the following criteria.

5.3.1    Contractors shall maintain the following documentation to verify the
         integrity of the standard solutions they purchase:

         •   Mass spectral identification confirmation of the solution,

         •   Purity confirmation of the solution, and

         •   Chromatographic and quantitative documentation that.the solution
             standard was QC checked according to the following section.

5.3.2    The Contractor shall purchase standards for which the quality is
         demonstrated statistically and analytically.  One way this may be
         demonstrated is to prepare and analyze three solutions, a high
         standard, a low standard, and a standard at the target concentration
          (see Sections 5.3.2.1 and 5.3.2.2).  The Contractor shall have
         documentation to demonstrate that the analytical results for the high
         standard and low standard are consistent with the difference in
         theoretical concentrations.  This is done by the Student's t-test in
         Section 5.3.2.4.  If this is achieved, the Contractor shall then

                                     E-19                              OLM04.2

-------
Exhibit E -- Section 5
Analytical Standards Requirements
         demonstrate that the concentration of the target standard lies midway
         between the concentrations of the low and high standards.   This is
         done by the Student's t-test in Section 5.3.2.5.  The standard is
         certified to be within 10% of the target concentration using the
         equations in Section 5.3.2.6.  If this procedure is used,  the
         Contractor shall document that the following have been achieved.

5.3.2.1     Two solutions of identical concentration shall be prepared
            independently from solutions.  An aliquot of the first solution
            shall be diluted to the intended concentration  (the "target
            standard").  One aliquot is taken from the second solution and
            diluted to a concentration 10% greater than the target standard.
            This is called the "high standard."  One further aliquot is taken
            from the second solution and diluted to a concentration 10,% less
            that the target standard.  This is called the "low standard."

5.3.2.2     Six replicate analyses of each standard  (a total of 18 analyses)
            shall be performed in the following sequence: low standard,
            target, high standard, low standard, target standard, high
            standard,  ...

5.3.2.3     The mean and variance of the six results for each solution shall
            be calculated.

            EQ. 2                              •
                                  Mean =
             EQ.  3

                                     ..£..-
                                             6 (MEAN)'
tn-
                           Variance =
             The values Y4 represent the results of the six analyses of each
             standard.   The  means of  the  low,  target,  and high standards  are
             designated M1( M2, and M3,  respectively.   The variances of the  low,
             target,  and high standards are designated V1( V2,  and V3,
             respectively.   Additionally,  a pooled variance, Vp, is calculated.
             EQ.  4
                                    V          V
                                         V,     3
                              V =  0.81    2   1.21
                               P           ^
             If the square root of Vp  is  less than 1% of M2, then M22/10,000
             shall be used as the value of Vp in all subsequent  calculations.

                                      E-20                              OLM04.2

-------
                                                        Exhibit E -- Section 5
                                             Analytical Standards Requirements
5.3.2.4     The test statistic shall be calculated.

            EQ. 5
                         Test Statistic =
                                           1.1
                                                 0.9
5.3.2.5
If the test statistic exceeds 2.13, then a 20% difference between
the high and low standards exists.  In such a case, the standards
are not acceptable.

The test statistic shall be calculated.
            EQ. 6
                        Test Statistic =
                                        k-
                                             1.8
                                                   2.2
5.3.2.6
If the test statistic exceeds 2.13, then the target standard
concentration has not been demonstrated to be the midway between
the high and low standards.   In such a case,  the standards are not
acceptable.

The 95% confidence intervals for the mean result of each standard
shall be calculated.
            EQ. 7
                   Interval for Low Standard =
                                    ± 2.13 J-J:
                                             6
      Interval for Target Standard =
                                                  ± 2-.13 J — -
                                                           6
                                     E-21
                                                                       OLM04.2

-------
Exhibit E -- Section 5
Analytical Standards Requirements
            EQ. 9

                   Interval for High Standard = M ± 2 .13 .1 —£
5.3.2.6.1      These intervals shall not overlap.  If overlap is observed, the
               ability to discriminate the 10% difference in concentrations
               has not been demonstrated.  In such a case, the standards are
               not acceptable.

5.3.2.6.2      In any event, the Contractor is responsible for the quality of
               the standards employed for analyses under this contract.

5.4   Requesting Standards From the EPA Standards Repository.  Solutions of
      analytical reference materials can be ordered from the U.S. EPA Chemical
      Standards Repository, depending on availability.  The Contractor can
      place an order for standards only after demonstrating.that these
      standards are not available from commercial vendors, either in solution
      or as a neat material.                                  .

5.5   Documentation of the Verification and Preparation of Chemical Standards.
      It is the responsibility of each Contractor to maintain the neces'sary
      documentation to show that the chemical standards they have used in the
      performance of CLP analysis conform to the requirements previously
      listed.

5.5.1    Weighing logbooks, calculations, chromatograms, mass spectra, etc.,
         whether produced by  the Contractor or purchased from chemical supply
         houses, shall be maintained by the Contractor  and may be subject to
         review during on-site laboratory evaluations.  In those cases where
         the documentation  is supportive of the  analytical results of data
         packages sent to the Agency, such documentation is to be kept on file
         by the Contractor  for a period of one year.

5.5.2    Upon  request by the  Technical Project Officer  or Administrative
         Project Officer, the Contractor shall submit their most recent
         previous year's documentation  (12 months)  for  the verification and
         preparation  of  chemical standards within  14 days of  the receipt  of
         request  to  the  recipients he/she designates.

5.5.3    The Agency may  generate a report discussing deficiencies in  the
         Contractor's  documentation  for  the verification and  preparation  of
         chemical  standards or may discuss  the deficiencies during an on-site
         laboratory  evaluation.  In  a detailed  letter to  the  Technical Project
         Officer  and the Administrative  Project  Officer,  the  Contractor shall
         address  the  deficiencies and the  subsequent  corrective  action
         implemented by the Contractor  to  correct  the deficiencies within 14
         days  of  receipt of the  report  or  the on-site  laboratory evaluation.
         An alternate delivery  schedule  may be proposed by  the  Contractor,  but
          it is the  sole  decision of  the  Agency,  represented either by the
         Technical  Project  Officer or Administrative  Project  Officer,  to
         approve  or  disapprove  the alternate  delivery schedule.   If  an

                                      E-22           '                   OLM04.2

-------
                                                        Exhibit E --  Section 5
                                             Analytical Standards Requirements


         alternate delivery schedule is proposed,  the Contractor shall
         describe in a letter to the Technical Project Officer, Administrative
         Project Officer, and the Contracting Officer why he/she is unable to
         meet the delivery schedule listed in this section.   The Technical
         Project Officer/Administrative Project Officer will not grant an
         extension for greater than 14 days for the Contractor's response
         letter to the standards documentation report.  The  Contractor shall
         proceed and not assume that an extension will be granted until so
         notified by the Technical Project Officer and/or Administrative
         Project Officer.

5.5.4    If new SOPs are required to be written or SOPs are  required to be
         amended because of deficiencies and the subsequent  corrective action
         implemented by the Contractor, the Contractor shall write/amend and
         submit the SOPs per the requirements listed in Section 4.

5.6   Corrective Actions.  If the Contractor fails to adhere to the
      requirements listed in Section 5, the Contractor may expect, but the
      Agency is not limited to, the following actions:  reduction of number of
      samples sent under the contract, suspension of sample  shipment to
      Contractor, a GC/MS and/or GC/EC tape audit, a data package audit,  an
      on-site laboratory evaluation, a remedial laboratory evaluation sample,
      and/or contract sanctions.
                                     E-23                              OLM04.2

-------
Exhibit E -- Section 6
Determination of Method Equivalency for Alternate Extraction Procedures
6.0   DETERMINATION OF METHOD EQUIVALENCY FOR ALTERNATIVE EXTRACTION
      PROCEDURES

      If the Contractor wishes to use one or both of the alternative
      extraction procedures described in Section 1.3 Exhibits D SVOA and D
      PEST under Scope and Application, the Contractor must develop and
      implement SOPs for performing  the alternative extractions in accordance
      to Exhibit E Section 4.0.  In  addition, the Contractor shall maintain
      documentation, including raw data, to demonstrate the equivalence of the
      alternative extraction procedures to those specified in Sections
      10.1.4.4, Exhibit D SVOA and 10.1.5.3, Exhibit D PEST.  The required
      documentation for demonstrating extraction equivalence include an
      Initial Precision Recovery study as described below.

6.1   Initial Precision Recovery (IPR) Study

6.1.1    For the semivolatile fraction, the Contractor shall spike four  (4)
         solid samples  (e.g., anhydrous sodium  sulfate) with all  the target
         compounds at concentrations equal to three  (3) times 'the Contract
         Required Quantitation Limits  (CRQL) listed in Exhibit C  under
         semivolatiles. For pesticides/Aroclors, the Contractor shall spike
         four  (4) solid samples  with the single component  target  compounds and
         an additional four  (4)  solid samples with Aroclor 1254 only at
         concentrations equal to three  (3) times the CRQLs listed in Exhibit  C
         under pesticides/Aroclors.  Each sample must contain  the appropriate
         surrogates at the concentrations specified in Section 10.1.4.4,
         Exhibit D SVOA or Section 10.1.5.3.3,  Exhibit D PEST.

6.1.2    The Contractor shall achieve  the following recovery limits for  the
         matrix  spike compounds  in each of the  four replicates of the IPR
         study.

        |        Recovery Limits for_Matrix_S_p_ike Compounds

        |                      Semi vo1atiles

        |   Compound                              Recovery Limits

        |   Phenol                                     26-90

        |   2--Chlorophenol                            25-102

        |   N-Nitroso-di-n-propylamine               41-126

        \   4-Chloro-3-methylphenol                  26-103

        |   Acenaphthene                              31-137

        |   4-Nitrophenol                              11-114

        |   2,4-Dinitrotoluene                        28-89

        |   Pentachlorophenol                          17-109

        I   Pyrene                                    35-142
                                      E-24                              OLM04.2

-------
                                                        Exhibit E -- Section 6
     Determination of Method Equivalency for Alternative Extraction Procedures


       |                        Pesticides

       |   Compound                             Recovery Limits

       |   Gamma-BHC (Lihdane)                       46-127

       |   Heptachlor                                35-130

       |   Aldrin                                    34-132

       |   Dieldrin                                  31-134

       |   Endrin                                    42-139

       |   4,4'-DDT                                  23-134
6.1.3    The advisory limits for the mean percent recoveries (%R) of all other
         target compounds in the IPR study is 75% to 125% of the spiked
         amount.

6.1.4    The advisory limits for the % Relative Standard Deviation  (%RSD) of
         the IPR recoveries for each compound is 25% and shall not exceed 50%.

6.2   Analytical Protocol Required

6.2.1    The Contractor shall extract all IPR samples using Stf-846 Methods
         3541  (Revision 0, September 1994) and 3545  (Revision 0, December
         1996) modified where appropriate to achieve the requirements of this
         SOW  (i.e., CRQLs and all technical acceptance, criteria).  All
         modifications to the extraction procedure  (e.g., use of methylene
         chloride/acetone (1:1, v/v) for pesticide extraction) shall be
         adequately documented and submitted with the data package.

6.2.2    The Contractor shall follow the sample cleanup procedures described
         in Exhibit D SVOA for semivolatiles and Exhibit D PEST for
         pesticides/Aroclors.

6.2.3    The Contractor shall analyze the sample extracts for the IPR study
         following the procedures described in Section 10.6, Exhibit D SVOA
         for semivolatile compounds and Section 10.2,. Exhibit D PEST for
         pesticides/Aroclors compounds.

6.3   Quantitation Limits/Quality Control Requirements

6.3.1    The Contractor shall achieve the CRQLs specified in Exhibit C under
         semivolatiles and pesticides/Aroclors.

6.3.2    The Contractor shall follow all QC requirements outlined in Exhibit D
         SVOA and Exhibit D PEST including frequency of method blanks,
         instrument blanks,  instrument performance checks, initial and
         continuing calibrations or calibration verifications, internal
         standards, and surrogates.
                                     E-25                              OLM04.2

-------
Exhibit E -- Section 6
Determination of Method Equivalency for Alternate Extraction Procedures
6.3.3    All technical acceptance criteria for sample analysis, method blank,
         and instrument blank analyses'described in Exhibit D SVOA and Exhibit
         D PEST shall be met.

6.3.4    .All semivolatile surrogate recoveries shall be within the limits
         specified in Table  7, Exhibit D SVOA.

6.3.5    The advisory limits for the recovery of pesticide surrogates are 30%
         to 150%.

6.4   Data Deliverable Requirements

6.4.1    The Contractor shall submit data packages containing all
         documentation formatted as required in Exhibits B and H  (including,
         but not  limited to, SDG Narrative, appropriate summary forms, and raw
         data).   Each IPR replicate shall be reported as a separate sample
          (i.e., field sample) on Form  I.  All tuning data, initial calibration
         data, continuing calibration  data, and associated blanks with their
         raw data must be included in  the data package.  The Contractor  shall
         include  the source  of  the blank solid samples used for the IPR  study
         in the data deliverables.

6.4.2    The Contractor shall include  in the SDG Narrative a discussion  of any
         modifications to the extraction procedures and any problems
         encountered along with the resolutions.  The Contractor shall provide
         an explanation in the  SDG Narrative for any of the target compound
         recoveries that fall outside  the advisory limits.  A  summary of the
         IPR results with all calculations must also be included in the  SDG
         Narrative.

6.4.3    Simultaneous delivery  of the  complete Method Equivalency Data Package
         shall be made to  the following recipients:

         EPA:        Data Package will be delivered to  the laboratory's
                     Administrative Project Officer  (APO).

         SMO:         USEPA Contract Laboratory Program
                      Sample  Management Office  (SMO)1
                      2000 Edmund Hal ley Drive
                      Reston, VA 20191-3436

          QATS:       USEPA  Contract Laboratory Program
                      Quality Assurance Technical  Support  (QATS) Laboratory2
                      2700  Chandler Avenue, Building C
                      Las  Vegas, NV 89120
                      Attn: Data Audit  Staff
       1The Sample Management  Office  (SMO) is a contractor operated facility
 operating under the CLASS contract awarded and administered by the EPA.

       2The Quality  Assurance  Technical  Support  (QATS) Laboratory is a
 contractor operated facility operating under the QATS contract awarded and
 administered by the EPA.

                                      E-26                              OLM04.2

-------
                                                          Exhibit E -- Section 7
                                                   Contract Compliance Screening
|  7.0   CONTRACT COMPLIANCE SCREENING

|  7.1   Contract compliance screening  (CCS) is one aspect "of the Government's
        contractual right of inspection of analytical data. CCS examines the
        Contractor's adherence to the contract requirements based on the sample
        data package delivered to the Agency.

|  7.2   CCS is performed by the Sample Management Office  (SMO) under the
        direction of the Agency.  To assure a uniform review, a set of
        standardized procedures has been developed to evaluate the sample data
        package submitted by a Contractor against the technical and completeness
        requirements of the contract.  The government reserves the right to add
        and/or delete individual checks.  CCS results are used in conjunction
        with other information to measure overall Contractor performance and to
        take appropriate actions to correct deficiencies in performance.

|  7.3   CCS results are mailed to the Contractor and all other data recipients.
        The'Contractor has a period of time to correct deficiencies.  The
        Contractor shall send all corrections to the Regional client and SMO.

|  7.4   The Agency may generate a CCS trend report which summarizes CCS results
        over a given period of time.  The Agency may send the CCS trend report
        or discuss the CCS trend report during an on-site laboratory evaluation.
        In a detailed letter to the Technical Project Officer and Administrative
       -Project Officer, the Contractor shall address the deficiencies, and the
       ^•'subsequent corrective action implemented by the Contractor to correct
       '-the deficiencies within 14 days of receipt of the report or the pn-site
       ''laboratory evaluation.  An alternate delivery schedule may be proposed
        by the Contractor, but it is the sole decision of the Agency,
        represented by the Technical Project Officer or Administrative Project
        Officer, to approve or disapprove the alternate delivery schedule.  If
        an alternate delivery schedule is proposed, the Contractor shall
        describe in a letter to the Technical Project Officer, Administrative
        Project Officer, and Contracting Officer why he/she is unable to meet
        the delivery schedule listed in this section.  The Technical Project
        Officer/Administrative Project Officer will not grant an extension for
        greater than 14 days for the Contractor's response to the CCS trend
        report.  The Contractor shall proceed and not assume that an extension
        will be granted until so notified by the Technical Project Officer
        and/or Administrative Project Officer.

|  7.5   If new SOPs are required to be written or SOPs are required to be
        amended because of the deficiencies and the subsequent corrective action
        implemented by the Contractor, the Contractor shall write/amend and
        submit the SOPs per the requirements listed in Section 4.

        If the Contractor fails to adhere to the requirements listed in Section
        7, the Contractor may expect, but the Agency is not limited to, the
        following actions:  reduction of number of samples sent under the
        contract, suspension of sample shipment to the Contractor, a GC/MS
        and/or GC/EC tape audit, a data package audit, an on-site laboratory
        evaluation, a remedial laboratory evaluation sample, and/or contract
        sanctions.
                                       E-27                              OLM04.2

-------
I  Exhibit E -- Section 8
  Regional Data Review

|  8.0   REGIONAL DATA REVIEW

|  8.1   Contractor data are generated to meet the specific needs of the EPA
        Regions. In order to verify the usability of data for the intended
        purpose, each Region reviews data from the perspective of the end user,
        based upon functional aspects of data quality.  General guidelines for
        data review have been developed jointly by the Regions and the National
        Program Office.  Each Region uses these guidelines as the basis for data
        evaluation.  Individual Regions may augment the basic guideline review
        process with additional review based on Region-specific or site-specific
        concerns.  Regional reviews, like the sites under investigation, vary
        based on the nature of the problems under investigation and the Regional
        response appropriate to the specific circumstances.

|  8.2   Regional data reviews, relating usability of the data to. a specific
        site, are part of the collective assessment process.  They complement
        the review done at the Sample Management Office, which is designed to
        identify contractual discrepancies, and the review done by the Program
        Office, which  is designed to evaluate Contractor and method performance.
        These individual evaluations are integrated into a collective review
        that is necessary for Program and Contractor administration and
        management and may be used to take appropriate action to correct
        deficiencies in the Contractor's performance.
                                        E-28                              OLM04.2

-------
                                                        Exhibit E -- Section 5
                                                           Proficiency Testing
9.0   PROFICIENCY TESTING

      As a means of measuring and evaluating both the Contractor's and the
      method's analytical performance, the Contractor must participate in
      EPA's Proficiency Testing Program.  EPA's Proficiency Testing Program
      involves the analysis of case specific Performance Evaluation (PE)
      samples and the participation in interlaboratory Quarterly Blind  (QB)
      Audits.  The Contractor's analytical PE samples and QB results will be
      used by EPA to assess and verify the Contractor's continuing ability to
      produce acceptable analytical data in accordance with the contractual
      requirements.

9.1   Performance Evaluation Samples

9.1.1    The Performance -Evaluation sample(s) maybe scheduled with the
         Contractor as frequently as on an SDG-by-SDG basis.  The PE samples
         may be sent either by the Regional Client or the National Program
         Office.  PE samples will assist EPA in monitoring Contractor
         performance.

9.1.2    PE samples will be provided as either single-blinds (recognizable as
         a PE sample but of unknown composition), or as double-blinds  (not
         recognizable as a PE sample and of unknown composition) '.   The
         Contractor will not be informed of either the analytes or the
         concentrations in the PE samples.

9.1.3    The Contractor may receive the PE samples as either full volume
         samples or ampulated/bottled concentrates from EPA or a designated
         EPA Contractor.  The PE samples shall come with instructions
         concerning the unique preparation procedures, if any,  required to
         reconstitute the PE samples  (i.e., the required dilution of the PE
         sample concentrate). .PE samples are to be extracted and/or analyzed
         with the rest of the routine samples in the SDG.  The Contractor
         shall prepare and analyze the PE sample using the procedure described
         in the sample preparation and method analysis sections of Exhibit D.
         All contract required QC shall also be met.  The PE sample results
         are to be submitted in the SDG deliverable package per normal
         reporting procedures detailed in Exhibit B.

9.1.4    In addition to PE sample preparation and analysis, the Contractor
         shall be responsible for correctly identifying and quantitating the
         analytes included in each PE sample.  When PE sample results are
         received by EPA, the PE sample results will be evaluated for correct
         analytical identification'and quantitation.  EPA will notify the
         Contractor of unacceptable performance.  EPA reserves the right to
         adjust the PE sample acceptance windows in order to compensate for
         any unanticipated difficulties with a particular PE sample.

9.1.5    The Contractor shall demonstrate acceptable analytical performance
         for both identification and quantitation of PE sample analytes.  For
         unacceptable PE sample performance, EPA may take, but is not limited
         to the following actions: reduce value or rejection of data for the
         samples, SDG, or Case impacted; Show Cause and/or Cure Notice;
         reduction in the number of samples shipped to the laboratory;

                                     E-29                              OLM04.2

-------
I  Exhibit E -- Section 3
|  Proficiency Testing
           suspension of sample shipment; an on-site laboratory inspection; a
           full data package audit; and/or require the laboratory to analyze a
           Remedial QB.

  9.2   Quarterly Blind Audits

  9.2.1    Quarterly Blind  (QB) Audits may be scheduled concurrently with all
           contract laboratories up to a frequency of four times a year.  A
           Quarterly Blind Audit is a unique analytical case containing only
           Performance Evaluation samples  (i.e., referred to as Quarterly Blind
            (QB) samples).  The QB samples will be scheduled by the National
           Program Office through the CLASS Contractor.  QB samples will assist
           EPA  in monitoring Contractor performance.

  9.2.2    QB samples will be provided as single-blinds  (recognizable as a PE
           sample but of unknown composition).   The Contractor will not be
           informed of either  the analytes or the concentrations in the PE
           samples.

  9.2.3    The  Contractor may  receive the QB samples as either full volume
           samples or ampulated/bottled concentrates from EPA or a designated
           EPA  Contractor.  The QB samples shall come with instructions
           concerning  the unique preparation procedures, if any, required to
           reconstitute  the QB samples  (i.e., the required dilution of  the QB
           sample  concentrate).  The Contractor  shall prepare and analyze the  QB
           samples using the procedure described in the sample preparation and
           method  analysis  sections of Exhibit D.  All contract required QC
           shall also be met.  The QB sample results are to be submitted in  the
           SDG  deliverable package per normal reporting procedures detailed  in
           Exhibit B.

  9.2.4    In addition  to QB sample preparation  and analysis, the Contractor
           shall be responsible for correctly identifying and quantitating  the
           analytes included in each QB sample.  When QB sample results are
           received by EPA,  the QB sample  results will be scored for  correct
           analytical  identification and quantitation.   The QB sample scoring
            will be provided to the Contractor via coded  evaluation sheets, by
           analyte.  EPA will  notify  the Contractor of unacceptable performance.
           EPA  reserves the right  to adjust  the  PE sample acceptance  windows in
            order to  compensate for any unanticipated difficulties with  a
           particular  PE sample.   The  Contractor's QB sample performance  will  be
            assessed  into one of  the following three categories:

   9.2.4.1    Acceptable, No Response  Required:  Score greater  than or equal  to
               90 percent.   The data meets  most  or all of the scoring  criteria.
              No -response is required.

   9.2.4.2  .  Acceptable,  Response Explaining Deficiencies Required:  Score
               greater than  or equal  to 75  percent, but  less  than  90 percent.
               Deficiencies  exist  in  the Contractor's performance.  Corrective
               action  response  required.
                                        E-30                              OLM04.2

-------
                                                        Exhibit E -- Section 9
                                                           Proficiency Testing
9.2.4.3     Unacceptable Performance, Response Explaining Deficiencies
            Required: Score less than 75 percent.  Deficiencies exist in the
            Contractor's performance to the extent that the National Program
            Office has determined that the Contractor has not demonstrated the
            capability to meet the contract requirements.  Corrective action
            response required.

9.2.5    In the case of Section 9.2.4.2 or 9.2.4.3, the Contractor shall
         describe the deficiency(ies) and the action (s) taken to correct the
         deficiency(ies) in a corrective action letter to the Administrative
         Project Officer, the Technical Project Officer, and the CLP Quality
         Assurance Coordinator within 14 days of receipt of notification from
         the Agency.                                          . •      •

9.2.5.1     An alternate delivery schedule for the corrective action letter
            may be proposed by the Contractor, but it is the sole decision of
            the Agency, represented either by the Technical Project Officer or
            Administrative Project Officer, to approve or disapprove the
            alternate delivery schedule.  If an alternate delivery schedule is
            proposed, the Contractor shall describe in a letter to the
            Technical Project Officer, Administrative Project Officer, and
            Contracting Officer why the laboratory is unable to meet the
            original delivery schedule listed in Section 9.2.5.  The Technical
            Project Officer/Administrative Project Officer will not grant an
            extension for greater than 14 days for the Contractor's corrective
            action letter.  The Contractor shall proceed and hot assume that
     -~      an extension will be granted until so notified by the Technical
     •?"      Project Officer or Administrative Project Officer.

9.2.6    In the case of Section 9.2.4.2 or 9.2.4.3, if new SOPs are required
         to be written, or if existing SOPs are required to be rewritten or
         amended because of deficiencies and subsequent corrective action
         implemented by the Contractor, the Contractor shall write/amend the
         SOPs per the requirements listed in Exhibit E, Section 4.

9.2.7    For unacceptable QB sample performance  (Section 9.2.4.3), the EPA may
         take, but is not limited to the following actions: reduction in the
         number of samples shipped to the laboratory; suspension of sample
         shipment; an on-site laboratory inspection; a full data package
         audit; and/or require the laboratory to analyze a Remedial QB sample;
         and/or contract sanctions.

9.2.8    A Remedial QB Audit is a unique analytical case containing only QB
         samples.  A Remedial QB Audit may be scheduled by the National
         Program Office with the Contractor(s) for any of the following
         reasons: unacceptable PE sample performance, unacceptable QB sample
         performance, and/or major change in the laboratory (e.g., relocation,
         new owner, or high turn-over of key personnel).  Sections 9.2.2
         through 9.2.7 apply to the Remedial QB Audit process.
                                     E-31                              OLM04.2

-------
I  Exhibit E -- Section
|  Proficiency Testing
  9.2.9    If the Contractor fails to adhere to the requirements listed in
           Section 9, the Contractor may expect, but the Agency is not limited
           to, •- the following actions: reduction in the number of samples sent
           under the contract; suspension of sample shipment to the Contractor;
           a full data package audit; an on-site laboratory inspection; a
           Remedial QB sample; and/or contract sanctions.
                                        E-32                              OLM04.2

-------
I                                                         Exhibit E -- Section 10
j                                                     GC/MS and GC/EC Tape Audits


|  10.0  GC/MS AND GC/EC TAPE AUDITS

|  10.1  Overview.  Periodically, the Agency requests the GC/MS and GC/EC
        magnetic tapes from Contractors for a specific Case iri order to
        accomplish tape audits.  Generally, tape submissions and audits are
        requested for the following reasons.

        •   Program overview,

        •   Indication of data quality problems,

        •   Support for on-site audits, and

        •   Specific Regional requests.

|  10.1.1   Depending upon the reason for an audit, the tapes from a recent Case,
           a specific Case, or a laboratory evaluation sample may be requested.
           Tape audits provide a mechanism to assess adherence to contractual
           requirements and to ensure the consistency of data reported on the
|           hardcopy/electronic deliverables with that generated on the GC/MS and
|           GC/EC tapes.  This function provides external monitoring of Program
           QC requirements and checks adherence of the Contractor to internal QA
           procedures.  In addition, tape audits enable the Agency to evaluate
           the utility, precision, and accuracy of the analytical methods.

|  10.1.2   The Contractor shall store all raw and processed GC/MS and GC/EC data
           on magnetic tape, in appropriate instrument manufacturer's format,
|           uncompressed, and with no security codes.  This tape shall include
|           data for samples, all QC samples, blanks, matrix spikes, matrix spike
           duplicates, initial calibrations, continuing calibrations,
|           calibration verification standards, including resolution check
|           samples and performance evaluation mixtures, GPC single component and
|           multicomponent and Florisil cartridge check samples and associated
|           calibrations, and instrument performance check solutions  (BFB and
           DFTPP) as well as all Contractor-generated spectral libraries and
           quantitation reports required to generate the data package.  The
           Contractor shall maintain a written reference logbook of tape files
           of the EPA sample number, calibration data, standards, blanks, matrix
           spikes, and matrix spike duplicates.  The logbook shall include EPA
           sample numbers and standard and blank IDs, identified by Case and
           Sample Delivery Group.

|  10.1.3   The Contractor is required to retain the GC/MS and GC/EC tapes for
           365 days after submission of the reconciled Complete SDG File.  When
|           submitting GC/MS and GC/EC tapes to the Agency, the following
           materials shall be delivered in response to the request.

|  10.1.3.1    All associated raw data files for samples, all QC samples, blanks,
              matrix spikes, matrix spike duplicates, initial calibrations,
|              continuing calibrations, calibration verification-standards,
|              including resolution check samples and performance evaluation
|              mixtures, GPC single component and multicomponent Florisil
|              cartridge check samples and associated calibrations, and
              instrument performance check solutions  (BFB and DFTPP).

                                       E-33                              OLM04.2

-------
I  Exhibit E -- Section 10
I  GC/MS and GC/EC Tape Audits
|  10.1.3.2    All processed data files and quantitation output files associated
|              with the raw data files described in Section 10.1.3.1.

|  10.1.3.3    All associated identifications and calculation files  (method
|              files) used to generate the data submitted in the data package.

|  10.1.3.4    All Contractor-generated mass spectral library files  (NIST/EPA/NIH
              and/or Wiley, or equivalent, library not required).

|  10.1.3.5    A copy of the Contractor's written reference logbook  relating  tape
              files to EPA sample number, calibration data, standards, blanks,
              matrix spikes, and matrix spike duplicates.  The logbook shall
              include EPA sample numbers and lab file identifiers for all
              samples, blanks, and  standards, identified by Case and Sample
              Delivery Group.

|  10.1.3.6    A directory of all files on each tape, including all
|              subdirectories and the  files contained therein.
I
|  10.1.3.7    A copy of the completed sample data package.

|  10.1.3.8    A statement attesting to the completeness of the GC/MS and GC/EC
              data  tape submission, signed and dated by the Contractor's
              laboratory manager.   The Contractor shall also provide a statement
              attesting that the data reported have not been altered in  any  way.
              These statements shall  be part of a cover sheet that  includes  the
              following information relevant to the data tape submission-

              •   Contractor name,

              •   Date of submission,

              •   Case number,

              •   SDG number,

|              •   GC/MS and GC/EC make and model number,
                                                                          <
              •   Software  version,

              •   Disk drive  type  (e.g.,  CDC,  PRIAM, etc.),

              •   File transfer  method  (e.g.,  DSD,  DTD,  FTP, Aquarius,  etc.),
                  and

|              •   Data System Computer,

|              •   System  Operating Software,

|              •   Data System Network,

|            .  •    Tape Backup Software,

|              •    Tape Backup Hardware,

                                        E-34                               OLM04.2

-------
                                                         Exhibit E -- Section 10
                                                     GC/MS and GC/EC Tape Audits
|              •   Data Analysis Software,

|              •   Fraction, and

|              •   Volume of data  (in Mb) backed up on each tape

              •   Names and telephone numbers of two Contractor contacts for
                  further information regarding the submission.

|  10.2  Submission of the GC/MS and GC/EC Tapes.  Upon request of the
        Administrative Project Officer, the Contractor shall send the required
|        GC/MS and/or GC/EC tapes and all necessary documentation to the EPA
        designated recipient  (e.g., QATS) within seven. (7) days of notification.
        An alternate delivery schedule may be proposed by the Contractor, but it
        is the sole decision of the Agency, represented either by the Technical
        Project Officer or Administrative Project Officer, to approve or
        disapprove the alternate delivery schedule.  If an alternate delivery
        schedule is proposed, the Contractor shall describe in a letter to the
        Technical Project Officer, Administrative Project Officer, and the
        Contracting Officer why he/she is unable to meet the delivery schedule
        listed in this section.  The Technical Project Officer/Administrative
        Project Officer will not grant an extension for greater than seven days
|        for submission of the GC/MS and/or GC/EC tape.  The Contractor shall
        "proceed and not assume that an extension will be granted until so
       .^notified by the Technical Project Officer and/or Administrative Project
       a Officer.

|        NOTE:  The GC/MS and GC/EC tapes shall be shipped according to the
        procedures in Exhibit F.

|  10.3  "Responding to the GC/MS and GC/EC Tape Audit Report.  After completion
|        of the GC/MS and GC/EC tape audit, the Agency may send a copy of the
I        GC/MS and GC/EC tape audit report to the Contractor or may discuss the
|        GC/MS and GC/EC tape audit report at an on-site laboratory evaluation.
        In a detailed letter to the Technical Project Officer and Administrative
        Project Officer, the Contractor shall discuss the corrective actions
|        implemented to resolve the deficiencies listed in the GC/MS and GC/EC
        tape audit report within 14 days of receipt of the report.

|  10.3.1   An alternate delivery schedule may be proposed by the Contractor, but
           it is the sole decision of the Agency, represented either by the
           Technical Project Officer or Administrative Project Officer, to
           approve or disapprove the alternate delivery schedule. ""If an
           alternate delivery schedule is proposed, the Contractor shall
           describe in a letter to the Technical Project Officer, Administrative
           Project Officer, and the Contracting Officer why he/she is unable to
           meet the delivery schedule listed in this section.  The Technical
           Project Officer/Administrative Project Officer will not grant an
           extension for greater than 14 days for the Contractor's response
|           letter to the GC/MS and GC/EC tape report.  The Contractor shall
           proceed and not assume Jthat an extension will be granted until so
           notified by the Technical Project Officer and/or Administrative
           Project Officer.                                 •
                                       E-35                              OLM04.2

-------
I  Exhibit E -- Section 10
|  GC/MS and GC/EC Tape Audits


|  10.3.2   If new SOPs are required to be written or SOPs are required to be
           amended because of the deficiencies and the subsequent corrective
           action implemented by the Contractor, the Contractor shall
           write/amend and submit the SOPs per the requirements listed in
           Section 4.

|  10.3.3   Maintenance of the Magnetic Tape Storage Device

|  10.3.3.1    The Contractor shall certify that the tape head alignment on the
              magnetic tape storage device is in compliance with the ANSI
              standards for nine track magnetic tapes.  If the Contractor does
              not have documentation of alignment within the last 12 months, the
              Contractor must perform or have performed the manufacturer's
              documented head alignment procedure within 60 days of contract
              award.  This is generally performed with a "skew" tape, certified
              to be  in conformance with ANSI standards.  The alignment must be
              performed by qualified personnel.  The tape head alignment must be
              performed at a minimum once every 12 months or when there is
              evidence that the tape head may be out of alignment.

|  10.3.3.2    The tape system, including recording head, must be in conformance
              with the manufacturer's physical and electrical standards.
              Alignment of the remaining components of the tape system such as
              the retracting arms, must be performed at intervals not to exceed
              24 months.  If the Contractor cannot demonstrate that the
              remaining components of the tape system are in alignment, then the
              Contractor must perform or have performed the manufacturer's
              recommended alignment procedure.

|  10.3.4   Record of Maintenance'of  the Magnetic  Tape Storage Device

           Documentation of maintenance, alignment, and repair procedures must
           be kept in an instrument maintenance log book for each tape device
           and  data  system.  Also  include any  local area network components that
           provide a means  for the transmission of data to or  from  the
           instrument data  system  and  the tape  system.  Maintenance entries must
           include serial number,  property  number (if applicable),  data and time
           of repair, name  of person performing maintenance, problem
           description, problem  resolution,  date  and  time of  failure  (if
           applicable), and date and  time placed  back in service.   Copies of
           repairs shall be kept in  the maintenance documentation.
           Documentation of  1) data  system,  and 2) tape system maintenance  and
           alignments,  for  the  last  24 months  must be made  available  upon
|   "       written request  of ""the  Technical Project Officer  or Administrative
|          Project Officer  or during  a laboratory on-site evaluation.  The
j          Contractor shall always submit  a GC/MS and GC/EC tape  from a tape
            system in conformance with the manufacturer's physical  and electrical
            standards and  alignment according to manufacturer's procedures.

|  10.4  Corrective Actions.   If  the  Contractor fails  to adhere  to  the
|        requirements listed in  Section 10,  the Contractor may expect,  but  the
         Agency  is not limited to,  the  following actions:  reduction in the
         number  of  samples  sent under the  contract,  suspension of  sample  shipment
|        to the  Contractor,  an on-site laboratory  evaluation,  a GC/MS  and/or

                                        E-36                              OLM04.2

-------
                                                 Exhibit E -- Section 10
                                             GC/MS and GC/EC Tape Audits
GC/EC tape audit, a data package audit,  a remedial laboratory evaluation
sample, and/or contract sanctions.
                               E-37                              OLM04.2

-------
  Exhibit E -'- Section 11
  Data Package Audits
I  11.0  DATA PACKAGE AUDITS

|  11.1  Overview.  Data package audits are performed by the Agency for program
        overview and specific Regional concerns and to assess the technical
        quality of the data and evaluate overall Contractor performance.  They
        provide the Agency with an in-depth inspection and evaluation of the
        Case data package with regard to achieving QA/QC acceptability.  Data
        packages are periodically selected from recently received Cases.  They
        are evaluated for the technical quality of hardcopy raw data, quality
        assurance, and adherence to contractual requirements.  A thorough review
        of the raw data is completed, including:  a check of instrument
        printouts, quantitation reports, chromatograms, spectra, library
        searches and other documentation for deviations from the contractual
        requirements, a check for transcription and calculation errors, a review
        of the qualifications of the Contractor personnel involved with the
|        Case, and a review of the latest version of all SOPs on file.
        Standardized procedures have been established to assure uniformity of
        the auditing process.

|  11.2  Responding to the Data Package Audit Report.  After completing  the data
        package audit, the Agency may send a copy of the data package audit
        report to the Contractor or may discuss the data package audit  report.at
        an on-site laboratory evaluation.  In a detailed letter to the  Technical
        Project Officer and  the Administrative Project Officer, the  Contractor
        shall discuss the corrective actions implemented to resolve  the
        deficiencies listed  in the data package audit report within  14  days of
        receipt of the report.

|  11.2.1   An alternate delivery schedule may be proposed by the Contractor, but
           it is  the sole decision of the Agency, represented either by the
           Technical Project Officer or Administrative Project Officer, to
           approve or disapprove the alternate delivery schedule.  If an
           alternate delivery schedule  is proposed, the Contractor shall
           describe  in a  letter to the Technical Project Officer, Administrative
           Project Officer,  and the  Contracting Officer, why he/she  is  unable to
           meet  the  delivery schedule listed in this section.  The Technical
           Project Officer/Administrative Project Officer will not grant an
           extension for  greater than 14 days for the Contractor's response
           letter to the  data package report.  The Contractor shall  proceed and
           not assume that an extension will be granted until so notified by the
           Technical Project Officer and/or Administrative Project Officer.

|  11.2.2   If new SOPs are required  to be written or SOPs are required  to be
           amended because of the deficiencies and the subsequent corrective
           action implemented by the Contractor, the Contractor shall
           write/amend and submit the SOPs per the requirements listed  in
           Section 4.

|  11.3  Corrective Actions.   If the  Contractor fails to adhere  to the
j        requirements listed  in Section  11, the Contractor may expect, but the
        Agency  is not  limited to, the following actions:  reduction  in  the
        numbers  of samples sent under the contract, suspension  of sample
        shipment to  the  Contractor,  an  on-site  laboratory evaluation, a GC/MS
|        and/or  GC/EC tape audit, a data package audit, a remedial laboratory
        evaluation sample, and/or contract sanctions.


                                        E-38                              OLM04.2

-------
I                                                         Exhibit E -- Section 12
                                                  On-Site Laboratory Evaluations


|  12.0  ON-SITE LABORATORY EVALUATIONS

|  12.1  Overview.  At a frequency dictated by a Contractor's performance, the
        Administrative Project Officer, Technical Project Officer, or the
        Contracting Officer will conduct an on-site laboratory evaluation.  On-
        site laboratory evaluations are carried out to monitor the Contractor's
        ability to meet selected terms and conditions specified in the contract.
        The evaluation process incorporates two separate categories:  a quality
        assurance evaluation and an evidentiary audit.

|  12.2  Quality Assurance On-Site Evaluation.  Quality assurance evaluators
        inspect the Contractor's facilities to verify the adequacy and
        maintenance of instrumentation, the continuity of personnel meeting
        experience or education requirements, and the acceptable performance of
        analytical and QC procedures.

|  12.2.1 •  The Contractor shall expect that items to be monitored will include,
           but not be limited to, the following items:

           •   Size and appearance of the facility,

           •   Quantity, age, availability, scheduled maintenance, and
               performance of instrumentation,

           •   Availability, appropriateness, and utilization of the QAP and
               SOPs,

           •   Staff qualifications and experience, and personnel training
               programs,

           •   Reagents, standards, and sample storage facilities,

           •   Standard preparation logbooks and raw data,

           •   Bench sheets and analytical logbook maintenance and review, and

           •   Review of the Contractor's sample analysis/data package
               inspection/data management procedures.

|  12.2.2   Prior to an on-site evaluation, various documentation pertaining to
           performance of the specific Contractor is  integrated in a profile
           package for discussion during the evaluation.  Items that may be
           included are previous on-site reports, laboratory evaluation sample
|           scores, Regional review of data, Regional QA materials, GC/MS and
|           GC/EC tape audit reports, data audit reports, results of CCS, and
           data- trend reports.

|  12.3  Evidentiary Audit.  Evidence auditors conduct an on-site laboratory
        evaluation to determine if Contractor policies and procedures are in
        place to satisfy evidence handling requirements as stated in Exhibit F.
        The evidence audit comprises a procedural audit, an audit of written
        SOPs, and an audit of analytical project file documentation.
                                       E-39                              OLM04.2.

-------
I  Exhibit E -- Section 12
  On-Site Laboratory Evaluations


|  12.3.1   Procedural Audit.  The procedural audit consists of review and
           examination of actual standard operating procedures and accompanying
           documentation for. the following Contractor operations:  sample
           receiving, sample storage, sample identification, sample security,
           sample tracking  (from receipt to completion of analysis), and
           analytical project file organization and assembly.

|  12.3.2   Written SOPs Audit.  The written SOPs audit consists  of review and
           examination of the written SOPs to determine  if they  are accurate and
           complete for the  following Contractor operations:  sample receiving,
           sample storage, sample identification, sample security, sample
           tracking  (from receipt to completion of analysis), and analytical
           project file organization and assembly.

|  12.3.3   Analytical Project File Evidence Audit.  The  analytical project  file
           evidence audit consists of review and examination of  the analytical
           project file documentation.  The auditors review the  files to
           determine:

           •    The accuracy  of  the document inventory,

           •    The completeness of the  file,

           •    The adequacy  and accuracy of the document numbering system,

           •    Traceability  of  sample activity,

           •    Identification of activity  recorded on  the documents, and

           •    Error  correction methods.

|  12.4  Discussion of the On-Site Team's Findings.  During  the  debriefing,  the
         auditors present  their  findings and recommendations  for corrective
         actions necessary to the Contractor personnel.

|  12.5  Corrective Action Reports for Follow-Through  to  Quality Assurance .and
         Evidentiary Audit Reports.   Following  an  on-site laboratory  evaluation,
         quality assurance and/or evidentiary audit reports  which discuss
         deficiencies  found  during the on-site  evaluation may be sent to the
         Contractor.   In a detailed  letter,  the Contractor  shall discuss the
         corrective  actions  implemented  to  resolve the  deficiencies  discussed
         during the  on-site  evaluation and  discussed  in the  report(s) to the
         Technical Project Officer and the  Administrative Project Officer within
         14  days of  receipt  of the report.

|  12.5.1  An alternate delivery schedule  may  be  proposed  by the Contractor, but
            it is the  sole decision  of the  Agency,  represented  either by the
            Technical  Project Officer or Administrative Project  Officer, to
            approve  or disapprove  the alternate delivery schedule.   If  an
            alternate  delivery schedule  is  proposed,  the Contractor shall
            describe in a letter to  the  Technical  Project Officer,  Administrative
            Project  Officer,  and the Contracting Officer why he/she is  unable to
            meet the delivery schedule  listed in this  section.   The Technical
            Project  Officer/Administrative  Project .Officer  will  not grant an

                                        E-40                              OLM04.2

-------
I                                                         Exhibit E -- Section 12
                                                  On-Site Laboratory Evaluations


           extension for greater than 14 days for the Contractor's response
           letter to the quality assurance and evidentiary audit report.  The
           Contractor shall proceed and not assume that an extension will be
           granted until so notified by the Technical Project Officer and/or
           Administrative Project Officer.

|  12.5.2   If new SOPs are required to be written or SOPs are required to be
           amended because of the deficiencies and the subsequent corrective
          .action implemented by the Contractor, the Contractor shall
           write/amend and submit the SOPs per the requirements listed in
           Section 4.

|  12.6  Corrective Actions.  If the Contractor fails to adhere to the
|        requirements listed in Section 12, the Contractor may expect, but the
        Agency is not limited to, the following actions:  reduction in the
        number of samples sent under the contract, suspension of sample shipment
|        to the Contractor, an on-site laboratory evaluation, a GC/MS and/or
|        GC/EC tape audit, a data package audit, a remedial laboratory evaluation
        sample, and/or contract sanctions.
                                       E-41                              OLM04.2

-------
  Exhibit E -- Section 13
  Quality Assurance and Data Trend Analysis


|  13.0  QUALITY ASSURANCE AND DATA TREND ANALYSIS

|  13.1  Data submitted by Contractors are subject to review from several
        aspects:  compliance with contract-required QC, usability, and full data
        package evaluation.  Problems resulting from any of these reviews may
|        determine the need for a GC/MS and GC/EC tape audit, an on-site
        laboratory evaluation and/or a remedial laboratory evaluation sample.
        In addition, QC prescribed in the methods provides information that is
        continually used by the Agency to assess sample data quality, Contractor
        data quality and Program data quality via data trend analysis.  Trend
        analysis is accomplished by entering data into a computerized database.
        Statistical reports that evaluate specific anomalies or disclose trends
        in many areas, including the following, are generated from this
        database:

        •   Surrogate spike recovery,                    .

        •   Laboratory evaluation sample results,

        •   Blanks,

        •   GC/MS instrument performance checks  (BFB and DFTPP),

        •   Initial and continuing calibration data, and

        •   Other QC and method parameters.

|  13.2  Program-wide statistical results are used to rank Contractors in order
        to observe  the relative performance of each Contractor  using a given
        protocol against its peers.  The reports are also used  to identify
        trends within Contractors.  The results of many of  these trend analyses
        are included in the overall evaluation of a Contractor's performance,
        and are reviewed to determine  if corrective action  or an on-site
        laboratory  evaluation may be required to ensure that the Contractor can
        meet  the QA/QC requirements of the contract.   Contractor performance
        over  time is monitored using these trend analysis techniques to detect
        departures  of Contractor output from required  or desired levels of
        quality control, and to provide an early warning of Contractor QA/QC
        problems which may not be apparent from the results of  an individual
        Case.

|  13.3  As a  further benefit to the Program, the database provides the
        information needed to establish performance-based criteria in updated
        analytical  protocols, where advisory criteria  have  been previously used.
        The vast empirical data set produced by Contractors is  carefully
        analyzed, with the results augmenting theoretical and research-based
        performance criteria.  The result  is a continuously monitored set of
        quality control  and performance criteria specifications of what is
        routinely achievable and  expected  of environmental  chemistry Contractors
        engaged in  mass production analysis of environmental samples.  This, in
        turn,  assists the  Agency  in meeting its objectives  of obtaining data of
        known and documented quality.
                                        E.-42                              OLM04.2

-------
I                                                         Exhibit E -- Section 14
                                                                 Data Management


|  14.0  DATA MANAGEMENT

|  14.1  Data management procedures are defined as procedures specifying the
        acquisition or entry, update, correction, deletion, storage, and
        security of computer-readable data and files.  These procedures shall be
        in written form and contain a clear definition for all databases and
        files used to generate or resubmit deliverables.  Key areas of concern
        include system organization  (including personnel and security),
        documentation operations, traceability, and quality control.

|  14.2  Data manually entered from hardcopy shall be subject to quality control
        and the error rates estimated.  Systems shall prevent entry of incorrect
        or out-of-range data and alert data entry personnel of errors.  In
        addition, data entry error rates shall be estimated and recorded on a
        monthly basis by reentering a statistical sample of the data entered and
        calculating discrepancy rates by data element.

|  14.3  The record of changes in the form of corrections and updates to data
        originally generated, submitted, and/or resubmitted shall be documented
        to allow traceability of updates.  Documentation shall include the
        following for each change.

        •   Justification or rationale for the change.

        •   Initials of the person making the change(s).  Data changes shall be
            implemented and reviewed by a person or group independent of the
            source generating the deliverable.

        •   Documentation of changes shall be retained according to the schedule
            of the original deliverable.

        •   Resubmitted diskettes or other deliverables shall be reinspected as
            a part of the Contractor's internal inspection process prior to
            resubmission.  The entire deliverable, not just the changes, shall
            be inspected.

        •   The Contractor's laboratory manager shall approve changes to
            originally submitted deliverables.

        •   Documentation of data changes may be requested by Contractor
            auditors.

|  14.4  Life cycle management procedures shall be applied to computer software
        systems developed by the Contractor to be used to generate and edit
        contract deliverables.  Such systems shall be thoroughly tested and
        documented prior to utilization.

|  14.4.1   A software test and acceptance plan including test requirements, test
           results, and acceptance criteria shall be developed, followed, and
           available in written form.
                                       •E-43                              OLM04.2

-------
Exhibit E -- Section 14
Data Management
14.4.2   System changes shall not be made directly to production systems
         generating deliverables.  Changes shall be made first to a
         development system and tested prior to implementation.

14.4.3   Each version of the.production system will be given an identification
         number, date of installation, date of last operation, and archived.

14.4.4   System and operations documentation shall be developed and maintained
         for each system.  Documentation shall include a user's manual and an
         operations and maintenance manual.

14.4.5   This documentation shall be available for on-site review and/or upon
         written request by the Technical Project Officer or Administrative
         Project Officer.

14.5  Individual(s) responsible for the following functions shall be
      identified.

      •   System operation and maintenance, including documentation and
          training,

      •   Database integrity, including data entry, data updating and quality
          control, and

      •   Data  and system security, backup, and archiving.
                                      E-44                              OLM04.2

-------
                EXHIBIT F
   CHAIN-OF-CUSTODY, DOCUMENT CONTROL,
AND WRITTEN STANDARD OPERATING PROCEDURES
                   F-l                               OLM04.2

-------
             Exhibit F - Chain-of-Custody, Document Control,  and
                    Written  Standard  Operating  Procedures

                              Table of  Contents
Section                                                                   Page

1.0   INTRODUCTION  	  3

2.0   STANDARD OPERATING PROCEDURES 	  4
      2.1   Sample Receiving  	 	  4
      2.2   Sample Identification 	  5
      2.3   Sample Security 	  5
      2.4   Sample Storage  	  6
      2.5   Sample Tracking and Document Control	  .  6
      2.6   Computer-Resident Sample Data Control	 .  .  .  7
      2.7   Complete Sample Delivery Group File  (CSF) Organization and
            Assembly   	  7

3.0   WRITTEN STANDARD OPERATING PROCEDURES  (SOPS)   	 10
      3.1   Sample Receiving	10
      3.2   Sample Identification 	 11
      3.3   Sample Security	12
      3.4   Sample Storage	12
      3.5   Sample Tracking and Document Control  	 12
      3.6   Computer-Resident Sample Data Control 	 13
      3.7   CSF Organization and Assembly	14
                                      F-2                               OLM04.2

-------
                                                        Exhibit F -- Section 1
                                                                  Introduction
1.0   INTRODUCTION
1.1   A sample is physical evidence collected from a facility or from the
      environment.  Controlling evidence is an essential part of the hazardous
      waste investigation effort. "To ensure that the Environmental Protection
      Agency's (EPA) sample data and records supporting sample-related
      activities are admissible and have weight as evidence in future
      litigation, Contractors are required to maintain EPA samples under
      chain-of-custody and to account for all samples and supporting records
      of sample handling, preparation, and analysis.  Contractors shall
      maintain sample identity, sample custody, and all sample-related records
      according to the requirements in this exhibit.
1.2
The purposes of the evidence requirements include:
          Ensuring traceability of samples while in the possession of the
          Contractor.

          Ensuring custody of samples while in the possession of the
          Contractor.

          Ensuring the integrity of sample identity while in the possession of
          the Contractor.

          Ensuring sample-related activities are recorded on documents or in
          other formats for EPA sample receipt, storage, preparation,
          analysis, and disposal.

          Ensuring all laboratory records for each specified Sample Delivery
          Group will be accounted for when the project is completed.

          Ensuring that all laboratory records directly related to EPA samples
          are assembled and delivered to EPA or, prior to delivery, are
          available upon EPA's request.
                                      F-3                               OLM04.2

-------
Exhibit F -- Section 2
Standard Operating Procedures

2.0   STANDARD OPERATING PROCEDURES

      The Contractor shall implement the following standard operating
      procedures for sample receiving, sample identification, sample security,
      sample storage, sample tracking and document control, computer-resident
      sample data control, and complete sample delivery group file
      organization and assembly to ensure accountability of EPA sample
      chain-of-custody as well as control of all EPA sample-related records.

2.1   Sample Receiving

2.1..1    The Contractor shall designate a sample custodian responsible for
         receiving EPA samples.

2.1.2    The Contractor shall designate a representative to receive EPA
         samples in the event that the sample custodian is not available.

2.1.3    Upon receipt, the condition of shipping containers and sample
         containers shall be inspected and recorded on Form DC-1 by the sample
         custodian or his/her representative.

2.1.4    Upon receipt, the condition of the custody seals  (intact/broken)
         shall be inspected and recorded on Form DC-1 by the sample custodian
         or his/her representative.

2.1.5    The sample custodian or his/her representative shall verify and
         record on Form DC-1 the presence or absence of the  following
         documents accompanying the sample shipment:

         •   Custody  seals,

         •   Chain-of-custody records,

         •   Traffic  reports or packing lists,

         •   Airbills or  airbill stickers, and

         •   Sample  tags.

2.1.6    The sample  custodian or his/her representative shall verify and
         record on Form DC-1 the agreement or disagreement of information
         recorded on all  documents  received with samples and information
         recorded on sample  containers.

2.1.7    The sample  custodian or his/her representative shall record the
         following information on  Form DC-1 as  samples are received and
         inspected:

         •   Custody seal numbers  when present,

         •   Airbill or airbill  sticker numbers,

         •   Sample  tags  listed/not listed on chain-of-custody  records,

          •   Cooler  temperature,
                                      F-4                               OLM04.2

-------
                                                        Exhibit F --  Section 2
                                                 Standard Operating Procedures

         •    Date of receipt,

         •    Time of receipt,

         •    EPA sample numbers,

         •    Sample tag numbers,

         •    Assigned laboratory numbers,

         •    Samples delivered by hand,  and

         •    Problems and discrepancies.

2.1.8    The sample custodian or his/her representative shall sign, date,  and
         record the time on all accompanying forms,  when applicable,  at the
         time of sample receipt (for example,  chain-of-custody records,
         traffic reports or packing lists,  and airbills).  Note: Initials are
         not acceptable.

2.1.9    The Contractor shall contact the Sample Management Office (SMO)  to
         resolve problems and discrepancies including, but not limited to,
         absent documents, conflicting information,  absent or broken custody
         seals, absent temperature indicator bottle, and unsatisfactory sample
         condition.(for example,  leaking sample container).

2.1.10   The Contractor shall record resolution of problems and discrepancies
         by SMO.

2.2   Sample Identification

2.2.1    The Contractor shall maintain the identity of EPA samples and
         prepared samples (including extracted samples, digested samples,  and
         distilled samples)  throughout the laboratory.

2.2.2    Each sample and sample preparation container shall be labeled with
         the SMO number or a unique laboratory sample identification number.

2.3   Sample Security

2.3.1    The Contractor shall demonstrate that EPA sample custody is
         maintained from receiving through retention or disposal.  A sample is
         in custody if:

         •    It is in your possession; or

         •    It is in your view after being in your possession; or

         •    It is locked in a secure area after being in your possession;  or

         •    It is in a designated secure area.   (Secure areas shall be
             accessible only to authorized personnel.)

2.3.2    The Contractor shall demonstrate security of designated secure areas.
                                     -P-5                               OLM04.2

-------
Exhibit F -- Section 2
Standard Operating Procedures

2.4   Sample Storage

      The Contractor shall designate storage areas for EPA samples and
      prepared samples.

2.5   Sample Tracking and Document Control

2.5.1    The Contractor shall record all activities performed on EPA samples.

2.5.2    Titles which identify the activities recorded shall be printed on
         each page of all laboratory documents.   (Activities include, but are
         not limited to, sample receipt, sample storage, sample preparation,
         and sample analysis.)  When a document is a record of analysis, the
         instrument type and parameter group  (for example, GC/MS-VOA) shall be
         included in the title.

2.5.3    When columns are used to organize information recorded on laboratory
         documents, the information recorded  in the columns shall be
         identified in a column heading.

2.5.4    Reviewers' signatures shall be identified on laboratory documents
         when reviews are conducted.  Note:   Individuals recording review
         comments on computer-generated raw data  are not required to be
         identified unless the written comments address data validity.

2.5.5    The laboratory name shall be identified  on preprinted laboratory
         documents.

2.5.6    Each laboratory document entry shall be  dated with the month/day/year
          (for example,  01/01/90) and signed  (or initialed) by the
         individual(s)  responsible for performing the recorded activity at  the
         time the activity is recorded.

2.S.I    Notations on  laboratory documents shall  be recorded in ink.

2.5.8    Corrections to laboratory documents  and  raw data  shall be made by
         drawing single lines through the errors  and entering the correct
         information.   Information shall not  be obliterated or rendered
         unreadable.   Corrections and additions to information shall be signed
          (or initialed) and  dated.

2.5.9    Unused portions of  laboratory documents  shall be  lined-out.

2.5.10   Pages in bound and  unbound  logbooks  shall be sequentially numbered.

2.5.11   Instrument-specific run  logs shall be maintained  to enable  the
         reconstruction of  run  sequences.

2.5.12   Logbook entries shall  be in chronological order.

2.5.13   Logbook entries shall  include  only one Sample Delivery Group  (SDG)
         per page,  except  in the  events where the SDGs  "share" QC samples  (for
         example,  instrument run  logs and extraction logs).

2.5.14   Information  inserted into  laboratory documents  shall be affixed
         permanently  in place.  The  individual responsible for inserting

                                      F-6                               OLM04.2

-------
                                                        Exhibit F -- Section 2
                                                 Standard Operating Procedures

         information shall sign and date across the insert and logbook page at
         the time information is inserted.

2.5.15   The Contractor shall document disposal or retention of EPA samples,
         remaining-portions of samples, and prepared samples.

2.6   Computer-Resident Sample Data Control

2.6.1    Contractor personnel responsible for original data entry shall be
         identified at the time of data input.

2.6.2    The Contractor shall make changes to electronic data in a manner
         which ensures that the original data entry is preserved, the editor
         is identified, and the revision date is recorded.

2.6.3    The Contractor shall routinely verify the accuracy of manually
         entered data, electronically entered data, and data acquired from
         instruments.

2.6.4    The Contractor shall routinely verify documents produced by the
         electronic data collection system to ensure accuracy of the
         information reported.

2.6.5    The Contractor shall ensure that the electronic data collection
         system is secure.

2.6.5.1     The electronic data collection system shall be maintained in a
            secure location.

2.6.5.2     Access to the electronic data collection system functions shall be
            limited to authorized personnel through utilization of software
            security techniques (for example, log-ons or restricted
            passwords).

2.6.5.3     Electronic data collection systems shall be protected from the
            introduction of external programs or software  (for example,
            viruses).

2.6.6    The Contractor shall designate archive storage areas for electronic
         data and  the software required to access the data.

2.6.7    The Contractor shall designate an individual responsible for
         maintaining archives of electronic data, including the software.

2.6.8    The Contractor shall maintain the archives of electronic data and
         necessary software in a secure location.   (Secure areas shall be
         accessible only to authorized personnel.)

2.7   Complete Sample Delivery Group File  (CSF) Organization and Assembly

2.7.1    The Contractor shall designate a document control officer responsible
         for the organization and assembly of the CSF.

2.7.2    The Contractor shall designate a representative responsible for the
         organization and assembly of the CSF in the event that the document
         control officer is not available.

                                      F-7                               OLM04.2

-------
Exhibit F -- Section 2
Standard Operating Procedures


2.7.3    The Contractor shall maintain documents relating to the CSF in a
         secure location.

2.7.4    All original laboratory forms and copies of SDG-related logbook pages
         shall be included in the CSF.

2.7.5    Copies of laboratory documents in the CSF shall be photocopied in a
         manner to provide complete and legible replicates.

2.7.6    Documents 'relevant to each SDG including, but not limited to, the
         following shall be included in the CSF:

         •  logbook pages,             •  records of failed or attempted analysis,
         •  benchsheets,               •  custody records,
         •  mass spectra,              •  sample tracking records,
         •  chromatograms,             • . raw data summaries,
         •  screening records,         •  computer printouts,
         •  preparation records,       •  correspondence,
         •  re-preparation records,    •  FAX originals,
         •  analytical records,        •  library search results, and
         •  re-analysis records,       •  other.

2.7.7    The document control officer or his/her  representative  shall ensure
         that  sample tags  are encased in clear plastic bags before placing
         them  in the CSF.                                             •

2.7.8  .  CSF documents  shall be organized and assembled on an  SDG-specific
         basis.

2.7.9    Original documents which  include information relating to more than
         one SDG  (for example,  chain-of-custody records,  traffic reports,
         calibration logs)  shall be  filed in the  CSF of the  lowest SDG number,
         and copies of  these originals shall be placed  in the  other CSF(s).
         The document control officer or his/her  representative  shall record
         the following  statement on  the copies  in dark  ink:

                                      COPY
                 ORIGINAL DOCUMENTS ARE INCLUDED IN CSF
                                                            Signature
                                                             Date

 2.7.10   All CSFs shall be submitted with a completed Form DC-2.   All
          resubmitted CSFs shall be submitted with a new or revised Form DC-2.

 2.7.11   Each item in the CSF and resubmitted CSFs shall be inventoried and
          assembled in the order specified on Form DC-2.   Each page of  the  CSF
          shall be stamped with a sequential number.  Page number ranges shall


                                      F-8                               OLM04.2

-------
                                                        Exhibit F -- Section 2
                                                 Standard Operating Procedures

         be recorded in the columns provided on Form DC-2.  Intentional gaps
         in the page numbering sequence shall be recorded in the "Comments"
         section on Form DC-2.  When inserting new or inadvertently omitted
         documents, the Contractor shall identify them with unique accountable
         numbers.  The unique accountable- numbers and' the locations of the
         documents shall be recorded in the "Other Records" section' on Form
         DC-2.

2.7.12   Before shipping each CSF, the document control officer or his/her
         representative shall verify the agreement of information recorded on
         all documentation and ensure that the information is consistent and
         the CSF is complete.-

2.7.13   The document control officer or his/her representative shall document
         the shipment of deliverable packages including what was sent, to
         whom, the date, and the carrier used.

2.7.14   Shipments of deliverable packages, including resubmittals, shall be
         sealed with custody seals by the document control officer or his/her
         representative in a manner such that opening the packages would break
         the seals.

2.7.15   Custody seals shall be signed and dated by the document control
         officer or his/her representative when sealing deliverable packages.
                                      F-9                               OLM04.2

-------
Exhibit F -- Section 3
Written Standard Operating Procedures

3.0   WRITTEN STANDARD OPERATING PROCEDURES (SOPS)

      The Contractor shall develop and implement the following written SOPs
      for sample receiving, sample identification, sample security, sample
      storage, sample tracking and document control, computer-resident sample
      data control, and CSF file organization and assembly to ensure
      accountability for EPA sample chain-of-custody and control of all EPA
      sample-related records.

3.1   Sample Receiving

3.1.1    The Contractor shall have written SOPs for sample receiving which
         accurately reflect the procedures used by the laboratory.

3.1.2    The written SOPs  for sample receiving shall ensure that the
         procedures listed below are in use at the laboratory.

3.1.2.1     The condition  of shipping containers  and sample containers are
            inspected and  recorded on Form DC-1 upon receipt by the sample
            custodian or his/her representative.

3.1.2.2     The condition  of custody seals are inspected and recorded on Form
            DC-1 upon receipt by the sample custodian or his/her
            representative.

3.1.2.3     The presence or absence of the following documents accompanying
            the sample  shipment is verified and recorded on Form DC-1 by the
            sample custodian or his/her representative:

            •   Custody seals,

            •   Chain-of-custody records,

            •   Traffic reports or packing lists,

            •   Airbills or airbill  stickers, and

            •   Sample  tags.

3.1.2.4     The agreement  or disagreement of  information recorded  on  shipping
            documents with information recorded on sample  containers  is
            verified  and  recorded on Form DC-1 by the  sample  custodian or
            his/her representative.

3.1.2.5     The following  information is recorded on Form  DC-1 by  the sample
            custodian or his/her  representative as samples are received and
            inspected:

            •   Custody seal  numbers when present,

            •   Airbill or airbill  sticker numbers,

            •   Sample  tag numbers  listed/not listed on chain-of-custody
                records,

             •   Cooler temperature,

                                      F-10                              OLM04.2

-------
                                                          Exhibit F -- Section 3
                                           Written Standard Operating Procedures


              •   Date of receipt,

              •   Time of receipt,

              •   EPA sample numbers,

              •   Sample tag numbers,

              •   Assigned laboratory numbers,

              •   Samples delivered by hand, and

              •   Problems and discrepancies.

  3.1.2.6     All accompanying forms are signed, dated, and the time is
              recorded, when applicable, at the time of sample receipt (for
              example, chain-of-custody records, traffic reports or packing
              lists, and airbills) by the sample custodian or his/her
              representative.

  3.1.2.7     SMO is contacted to resolve problems and discrepancies including,
              but not limited- to, absent documents, conflicting information,
|              absent or broken custody seals,  absent temperature indicator
|              bottle, and unsatisfactory sample .condition (for example, leaking
              sample container).

  3.1.2.8     The resolution of problems and discrepancies by SMO is recorded.

  3.2   Sample Identification

  3.2.1    The Contractor shall have written SOPs for sample identification
           which accurately reflect the procedures used by the laboratory.

  3.2.2    The written SOPs for sample identification shall ensure that the
           procedures listed below are in use at the laboratory.

  3.2.2.1     The identity of EPA samples and prepared samples is maintained
              throughout the laboratory:

              •   When the Contractor assigns unique laboratory sample
                  identification numbers, the written SOPs shall include a
                  description of the procedure used to assign these numbers,
                                            t
              •"   When the Contractor uses prefixes or suffixes in addition to
                  laboratory sample identification numbers,  the written SOPs
                  shall include their definitions, and

              •   When the Contractor uses methods to uniquely identify
                  fractions/parameter groups and matrix type, the written SOPs
                  shall include a description of these methods.

  3.2.2.2     Each sample and sample preparation container is' labeled with the
              SMO number or a unique laboratory sample identification number.
                                       F-ll                              OLM04.2

-------
Exhibit F -- Section 3
Written Standard Operating Procedures


3.3   Sample Security

3.3.1    The Contractor shall have written SOPs for sample security which
         accurately reflect the procedures used by the laboratory.

3.3.2    The written.SOPs for sample security shall include the items listed
         below.

3.3.2.1     Procedures which ensure the following:

            •   Sample custody is maintained, and

            •   The security of designated secure areas is maintained.

3.3.2.2     A  list of authorized personnel who have access to locked storage
            areas.

3.4   Sample Storage    .                    /

3.4.1    The Contractor shall have written SOPs for sample storage which
         accurately reflect the procedures used by the laboratory.

3.4.2    The written SOPs for sample storage shall describe locations,
         contents, and identities of all  storage areas for EPA samples and
         prepared samples in the laboratory.

3.5   Sample Tracking and Document Control

3.5.1    The Contractor shall have written SOPs for sample tracking and
         document control which accurately reflect the procedures used by the
         laboratory.

3.5.2    The written SOPs for sample tracking and document control shall
         include  the items  listed below.

3.5.2.1     Examples of all laboratory documents used during sample receiving,
            sample storage, sample transfer, sample analyses, CSF organization
            and assembly, and sample retention or disposal.

3.5.2.2     Procedures which ensure the following:

            •   All  activities performed on  EPA samples are recorded;

            •   Titles which identify  the activities  recorded are printed on
               , each page of all  laboratory  documents;

            •   Information recorded in columns is  identified with column
                headings,-   •   .

            •   Reviewers'  signatures  are identified  on laboratory documents;

             •  • The  laboratory name  is included on preprinted laboratory
                documents;

             •   Laboratory  document  entries  are signed and dated  with the

                                     F-12                              OLM04.2

-------
                                                        Exhibit F -- Section 3
                                         Written Standard Operating Procedures

                month/day/year (for example,  01/01/90);

            •    Entries on all laboratory documents are recorded in ink;

            •    Corrections and additions to- laboratory documents are made by
                drawing single lines through the errors, entering the correct
                information, and initialing and dating the new information;

            •    Unused portions of laboratory documents are lined-out;

            •    Pages in bound and unbound logbooks are sequentially numbered;

            •    Instrument-specific run logs are maintained to enable the
                reconstruction of run sequences;

            •    Logbook entries are recorded in chronological order;

            •    Entries are recorded for only one SDG on a page, except in the
                events where SDGs "share" quality control (QC) samples  (for
                example, instrument run logs and extraction logs);

            •    Information inserted in laboratory documents is affixed
                permanently, signed, and dated across the insert; and

            •    The retention or disposal of EPA samples, remaining portions
                •of samples, and prepared samples is documented.

3.6   Computer-Resident Sample Data Control

3.6.1    The Contractor shall have written SOPs for computer-resident sample
         data control which accurately reflect the procedures used by the
         laboratory.

3.6.2    The written SOPs for computer-resident sample data control shall
         include the items listed below.

3.6.2.1     Procedures which ensure the following:

            •    Contractor personnel responsible for original data entry are
                identified;

            •    Changes to electronic data are made such that the original
                data entry is preserved, the editor is identified,  and the
                revision date is recorded;

            •    The accuracy of manually entered data, electronically entered
                data, and data acquired from instruments is verified;

            •    Report documents produced by the electronic data collection
                system are routinely verified to ensure the accuracy of the
                information reported;

           . •    'Electronic data collection system security is maintained; and

            •    Archives of electronic data and accompanying software are
                maintained in a secure location.

                                     F-13                              OLM04.2

-------
Exhibit F -- Section 3
Written Standard Operating Procedures

3.6.2.2     Descriptions of archive storage areas for the electronic data and
            the software required to access data archives.

3.6.2.3     A list of authorized personnel who have access to electronic data
            collection system functions and to archived data.

3.7   CSF Organization and Assembly

3.7.1    The Contractor shall have written SOPs for CSF organization and
         assembly which accurately reflect the procedures used by the
         laboratory.

3.7.2    The written SOPs for CSF organization and assembly shall ensure that
         the procedures listed below are in use at the laboratory.

         •   Documents relating to the CSF are maintained in a secure
             location.

         •   All original laboratory forms and copies of SDG-related logbook
             pages are included in the CSF.

         •   Laboratory documents are photocopied in a manner to provide
             complete and legible replicates.

         •   All documents  relevant to each SDG are included in the CSF.

         •   Sample  tags are encased in clear plastic bags by the document
             control officer or his/her representative before placing them in
             the CSF.

         •   The CSF is organized and assembled on an SDG-specific basis.

         •   Copies  are referenced to originals in the event that an original
             document contains information relating to more  than one SDG.

         •   Each  CSF is submitted with a  completed Form DC-2, and resubmitted
             CSFs  are submitted with a new or revised Form DC-2.

         •   Each  page of the CSF is stamped with a sequential number and the
             page  number ranges are recorded in the columns  provided on Form
             DC-2.

         •   Consistency and  completeness  of the CSF  is verified by the
             document control officer or his/her representative.

         •   Shipments of deliverable packages are documented by the document
             control officer  or his/her representative.

         •   Deliverable packages are shipped by the  document control officer
             or his/her  representative using custody  seals  in a manner  such
             that  opening  the packages would break the seals.

         •   Custody seals  are  signed and  dated by the document control
             officer or  his/her  representative before placing them on
             deliverable packages.
                                      F-14                              OLM04.2

-------
    EXHIBIT G
GLOSSARY OF TERMS
       G-l                               OLM04.2

-------
Exhibit G -- Glossary of Terras
ALIQUOT - a measured portion of a sample, or solution, taken for sample
preparation and/or analysis.

ANALYSIS DATE/TIME - the date and military time of the injection of the
sample, standard, or blank  into the GC/MS or GC system.

BAR GRAPH SPECTRUM - a plot of the mass-to-charge ratio  (m/e) versus relative
intensity of the ion current.

BLANK  - an analytical sample designed to assess specific sources of laboratory
contamination.  See individual types of Blanks:  Method Blank; Instrument
Blank, Storage Blank, and Sulfur Blank.

BREAKDOWN - a measure of the decomposition of certain analytes  (DDT and
Endrin) into by-products.

4-BROMOFLUOROBENZENE  (BFB)  - the compound chosen to establish mass spectral
instrument performance for  volatile  (VOA) analyses.  It is also used in the
VOA fraction as a system monitoring compound  (SMC).

CALIBRATION FACTOR  (CF)  - a measure of the gas chromatographic response of a
target analyte to the mass  injected.  The calibration factor is analogous to
the Relative Response Factor  (RRF) used  in the Volatile and Semivolatile
fractions.

CASE  - a finite, usually predetermined number of samples collected over a
given time period from a particular site.  Case numbers are assigned by the
Sample Management Office.   A Case consists of one  or more Sample Delivery
Groups.

CHARACTERIZATION  - a determination of the approximate concentration range of
compounds of interest used  to choose the appropriate analytical protocol.

CONCENTRATION LEVEL  (low or medium)  - characterization of soil  samples or
sample fractions  as  low  concentration or medium concentration is made on the
basis of the laboratory's preliminary screen, not  on the basis  of information
entered on the Traffic Report by the sampler.

CONTAMINATION -  a component of a sample  or an extract that is not
representative of the environmental  source of the  sample.  Contamination may
stem  from other  samples, sampling equipment, while in transit,  from laboratory
reagents, laboratory environment, or analytical instruments.

CONTINUING CALIBRATION  - analytical  standard  run every 12 hours to verify the
initial calibration  of  the  system.

CONTINUOUS LIQUID-LIQUID EXTRACTION  - used herein  synonymously  with the  terms
continuous extraction,  continuous liquid extraction,  and liquid extraction.
This  extraction  technique  involves boiling the extraction solvent in a flask
and condensing the  solvent  above the aqueous  sample.  The condensed solvent
drips through the sample,  extracting the compounds of  interest  from the
aqueous phase.
                                      G-2                               OLM04.2

-------
                                                Exhibit G -- Glossary of Terms
(CLASS) CONTRACT LABORATORY ANALYTICAL SERVICES SUPPORT - contract that
operates the Sample Management Office  (SMO) and is awarded and administered by
the EPA.

DATE -

      Hardcopy Reporting - MM/DD/YY - where MM = 01 for January, 02 for
      February, ... 12 for December; DD = 01 to 31; YY = 96, 97, 98, 99, etc.

      Electronic Reporting - MM/DD/YYYY - where MM = 01 for January, 02 for
      February, ... 12 for December; DD = 01 to 31; YYYY = 1996, 1997, 1998,
      1999, etc.

DAY - unless otherwise specified, day shall mean calendar day.
                                                                           >•
DECAFLUOROTRIPHENYLPHOSPHINE  (DFTPP) - compound chosen to establish mass
spectral instrument performance for semivolatile analysis.

EXTRACTABLE - a compound that can be partitioned into an organic solvent from
the sample matrix and is amenable to gas chromatography.  Extractables include
semivolatile  (BNA) and pesticide/Aroclor compounds.

EXTRACTED ION CURRENT PROFILE  (EICP) - a plot of ion abundance versus time (or
scan number) for ion(s) of specified mass(es).

FIELD SAMPLE - a portion of material to be analyzed that is contained in
single or multiple containers and identified by a unique sample number.

GAS CHROMATOGRAPH  (GC) - the  instrument used to separate analytes on a
stationary phase within a chromatographic column.  The analytes are volatized
directly from the sample  (VOA water and low-soil), volatized from the sample
extract  (VOA medium soil), or injected as extracts  (SVOA and PEST).  In VOA
and SVOA analysis, the compounds are detected by a Mass Spectrometer  (MS).  In
PEST analysis, the compounds  are detected by an Electron Capture  (EC)
detector. ' In the screening procedure  (all fractions), the Flame lonization
Detector  (FID) is used as the detector.

GEL PERMEATION CHROMATOGRAPHY  (GPC) - a size-exclusion chromatographic
technique that is used as a cleanup procedure for removing large organic
molecules, particularly naturally occurring macro-molecules such as lipids,
polymers, viruses, etc.

IN-HOUSE - at the Contractor's facility.

INITIAL CALIBRATION - analysis of analytical standards for a series of
different specified concentrations; used to define the linearity and dynamic
range of the response of the mass spectrometer or electron capture detector to
the target compounds.

INTEGRATION SCAN RANGE - the  scan number of the scan at the beginning of the
area of integration to the scan number at the end of the area of integration.
Performed in accordance with  Exhibit D VOA, Sections 11.2.1:9 and 11.2.1.10
and Exhibit D SVOA, Sections  11.2.1.2 and 11.2.1.3.
                                      G-3                              OLM04.2

-------
Exhibit G -- Glossary of Terms
INTEGRATION TIME RANGE - the retention time at the beginning of the area of
integration to the retention time at the end of the area of integration.

INTERNAL STANDARDS - compounds added to every standard, blank, matrix spike,
matrix spike duplicate, sample (for volatiles), and sample extract (for
semivolatiles) at a known concentration, prior to analysis.  Internal
standards are used as the basis for quantitation of the target compounds.

INSTRUMENT BLANK - a blank designed to determine the level of contamination
associated with the analytical instruments.

INSUFFICIENT QUANTITY - when there is not enough volume (water sample) or
weight  (soil/sediment) to perform any of the required operations:  sample
analysis or extraction, percent moisture, MS/MSD, etc. Exhibit D provides
guidance for addressing this situation.

LABORATORY - synonymous with Contractor as used herein.

m/z - Mass to charge ratio, synonymous with  "m/e".

MATRIX  - the predominant material of which the sample to be analyzed  is
composed. For the purpose of this SOW, a sample matrix is either water or
soil/sediment. Matrix is not synonymous with phase  (liquid or solid).

MATRIX  EFFECT -  in general, the effect of a particular matrix  (water or
soil/sediment) on the constituents with which it  contacts.  This is
particularly pronounced for clay particles which  may adsorb chemicals and
catalyze reactions.  Matrix effects may prevent extraction of target  analytes,
and may affect surrogate recoveries.   In addition, non-target analytes may be
extracted from the matrix causing interferences.

MATRIX  SPIKE  - aliquot of a matrix  (water or soil) fortified  (spiked) with
known quantities of specific compounds  and subjected to the entire analytical
procedure in  order to  indicate the appropriateness of  the method for  the
matrix  by measuring recovery.

MATRIX  SPIKE  DUPLICATE  - a second aliquot of the  same matrix  as the matrix
spike  (above) that is  spiked in order  to determine the precision of the
method.

METHOD  BLANK  - an  analytical control consisting of all reagents, internal
standards, and surrogate standards  (or SMCs  for VOA),  that is carried
throughout the entire  analytical procedure.   The  method blank is used to
define  the level of laboratory, background,  and reagent contamination.

NARRATIVE  (SDG Narrative)  - portion of the data package which includes
laboratory, contract,  Case, and sample number identification, and descriptive
documentation of any  problems  encountered  in processing the samples,  along
with corrective  action taken and problem resolution.   Complete  SDG Narrative
specifications are included  in Exhibit B.

PERCENT DIFFERENCE (%D)  - As used-in this  SOW and elsewhere to  compare  two
values, the percent difference indicates both the direction and the magnitude
of the  comparison,  i.e., the percent difference may  be either negative,
positive, or  zero.  (In contrast, see relative .percent  difference.)

                                      G-4                               OLM04.2

-------
                                                  Exhibit G -- Glossary of Terms


  PERCENT MOISTURE - an approximation of the amount of water in a soil/sediment
  sample made by drying an aliquot of the sample at 105 °C.   The percent
  moisture determined in this manner also includes contributions from all
  compounds that may volatilize at or below 105 °C,  including water.   Percent
  moisture may be determined from decanted samples and from samples that are not
  decanted.

  PERFORMANCE EVALUATION MIXTURE - a calibration solution of specific analytes
  used to evaluate both recovery and percent breakdown as measures of
  performance.

  PRIMARY QUANTISATION ION - a contract specified ion used to quantitate a
  target analyte.

  PROTOCOL - describes the exact procedures to be followed with respect to
  sample receipt and handling, analytical methods, data reporting and
  deliverables, and document control.  Used synonymously with Statement of Work
  (SOW).

  PURGE AND TRAP  (DEVICE) - analytical technique  (device) used to isolate
  volatile (purgeable) organics by stripping the compounds from water or soil by
  a stream of inert gas, trapping the compounds on an adsorbent such as a porous
  polymer trap, and thermally desorbing the trapped compounds onto the gas
  chromatographic column.

  PURGEABLES - volatile compounds.

|  QUALITY ASSURANCE TECHNICAL SUPPORT  (QATS) LABORATORY - a contractor operated
j  facility operated under the QATS contract, awarded and administered by the
|  EPA.

  REAGENT WATER - water in which an interferant is not observed at or above the
  minimum quantitation limit of the parameters of interest.

  RECONSTRUCTED ION CHROMATOGRAM  (RIC) - a mass spectral graphical
  representation of the separation achieved by a gas chromatograph; a plot of
  total ion current versus retention time.

  RELATIVE PERCENT DIFFERENCE  (RPD) - As used in this SOW and elsewhere to
  compare two values, the relative percent difference is based on the mean of
  the two values, and is reported as an absolute value, i.e., always expressed
  as a positive number or zero.  In contrast, see percent difference.

  RELATIVE RESPONSE FACTOR  (RRF) - a measure of the relative mass spectral
  response of an analyte compared to its internal standard.  Relative Response
  Factors are determined by analysis of standards and are used in the
  calculation of concentrations of analytes in samples.  RRF is determined by
  the following equation:


                                         Ax   Cis
                                 RRF  = —^- x ——
                                         A.     C
                                          IB   • X
        Where,

                                        G-5                              OLM04.2

-------
  Exhibit G -- Glossary of Terms
        A  = area of the characteristic ion measured
        C  = concentration,  or amount (mass)
        is = internal standard
        x  = analyte of interest

  RELATIVE RETENTION TIME (RRT) - the ratio of the retention time of a compound
  to that of a standard (such as an internal standard).
                                    RRT =
                                          RT
                                          RT
                                            is
        Where,

        RTC  = Retention time for the semivolatile target or surrogate compound
              in continuing calibration.
        RTis=  Retention time for the internal standard in calibration standard
              or in a sample.

|  REPRESENTATIVE - alternate or designee who has the knowledge and authority to
|  perform a specific task.

  RESOLUTION - also termed separation or percent resolution, the separation
  between peaks on a chromatogram, calculated by dividing the depth of the
  valley between the peaks by the peak height of the smaller peak being
  resolved,  multiplied by 100.
                         Am (or Height)
  For pesticide analysis the X-axis shall be displayed such that a data reviewer
  can calculate the  % Resolution.

  RESOLUTION CHECK MIXTURE  - a  solution of specific analytes used to determine
  resolution of adjacent peaks; used to assess  instrumental performance.

  RESPONSE  - or Instrumental Response: a measurement of the output of the GC
  detector  (MS, EC,  or  FID) in  which the intensity of the signal is
  proportionate to the  amount  (or concentration) detected.  Measured by peak
  area or peak height.
                                        G-6
OLM04.2

-------
                                                  Exhibit G -- Glossary of Terms


  RETENTION TIME (RT) - the time a target analyte is retained on a GC column
  before elution.  The identification of a target analyte is dependent on a
  target compound's retention time falling within the specified retention time
  window established for that compound.  Retention time is dependent on the
  nature of the column's stationary phase, column diameter, temperature, flow
  rate, and other parameters.

  SAMPLE DELIVERY GROUP (SDG) - a unit within a single Case that is used to
  identify a group of samples for delivery.  An SDG is a group of 20 or fewer
|  field samples within a Case, received over a period of up. to 7 calendar days.
  Data from all samples in an SDG are due concurrently.  A Sample Delivery Group
  is defined by one of the following, whichever occurs, first:

  •     All samples within a Case; or

|  •     Every set of 20 field samples  (excluding PE samples) within a Case; or

|  •     All samples received within  7 calendar days, excluding Sundays and
|        Government holidays.  However, PE samples received within a Case shall
|        be assigned to an SDG containing field samples for the Case.

|  In addition, all samples and sample fractions assigned to an SDG must have
|  been scheduled under the same contractual turnaround time.

  Samples may be assigned to Sample Delivery Groups by matrix (i.e., all soil
  samples in one SDG, all water samples in another), at the discretion of the
  laboratory.

|  SAMPLE MANAGEMENT OFFICE  (SMO)  - a contractor operated facility operated by
|  the CLASS contract, awarded and administered by the EPA.

  SAMPLE NUMBER  (EPA Sample Number)  - a unique identification number designated
  by EPA to each sample.  The EPA sample number appears on the sample Traffic
  Report which documents information on.that sample.

  SECONDARY QUANTITATION ION - contract specified ion(s) to be used in
  quantitation of target analytes when interferences prevent the use of the
  primary quantitation ion.

  SEMIVOLATILE COMPOUNDS - compounds amenable to analysis by extraction of the
  sample with an organic solvent.  Used synonymously with Base/Neutral/Acid
   (BNA) compounds.

  SOIL - used herein synonymously with soil/sediment and sediment.

  SONIC CELL DISRUPTOR  (SONICATOR) - a device that uses the energy from
  controlled ultrasound applications to mix, disperse, and dissolve organic
  materials from a given matrix.

  STANDARD ANALYSIS - an analytical determination made with known quantities of
  target compounds; used to determine response factors.

  STORAGE BLANK - reagent water  (two 40.0 mL aliquots) stored with samples in an
  SDG.  It is analyzed after all samples in that SDG have been analyzed; and is
  used to determine the level of contamination acquired during storage.

                                       G-7                              OLM04.2

-------
  Exhibit  G --  Glossary of Terms


  SULFUR BLANK  -  a modified method blank that is  prepared  only when some of  the
  samples  in a  batch are subjected to sulfur cleanup.   It  is  used to determine
  the level of  contamination associated with the  sulfur cleanup procedure.   When
  all of the samples are subjected to  sulfur cleanup,  then the method blank
  serves this purpose.   When none of the samples  are subjected to sulfur
  cleanup, no sulfur blank is required.

  SURROGATES (Surrogate Standard)  - for semivolatiles and  pesticides/Aroclors,
  compounds added to every blank,  sample, matrix  spike,  matrix spike duplicate,
  and standard; used to evaluate  analytical efficiency by  measuring recovery.
  Surrogates are brominated, fluorinated, or isotopically  labeled compounds  not
  expected to be detected in environmental media.

  SYSTEM MONITORING COMPOUNDS - compounds added to every blank, sample, matrix
  spike, matrix -spike duplicate,  and standard for volatile analysis, and used to
  evaluate the performance of the entire purge and trap-gas chromatograph-mass
  spectrometer system.   These compounds are brominated or  deuterated compounds
  not expected to be detected in environmental media.

  TARGET COMPOUND LIST  (TCL) - a list of compounds designated by the Statement
  of Work  (Exhibit C) for analysis.

  TENTATIVELY IDENTIFIED COMPOUNDS  (TIC) - compounds detected in samples, that
  are not target compounds, internal standards, system monitoring compounds, or
|  surrogates.  TICs must have peak areas or heights greater than 10% of the peak
|  areas or heights of nearest internal standard.   TICs must be subjected to mass
|  spectral library searches and be deemed acceptable by a mass spectral
|  interpretation specialist.

  TIME  - when required to record time on any deliverable item, time shall be
  expressed as Military Time, i.e., a 24-hour clock.

  TRAFFIC REPORT  (TR) - an EPA sample identification form filled out by the
  sampler, which accompanies the sample during shipment to the laboratory and
  which documents sample condition and receipt by the laboratory.

  TWELVE-HOUR TIME PERIOD - The twelve'(12) hour time period for GC/MS system
  instrument performance check, standards calibration  (initial or continuing
  calibration), and method blank analysis begins at the moment of injection of
  the DFTPP or BFB analysis that the laboratory submits as documentation of
  instrument performance.  The time period ends after 12 hours have elapsed
  according to the system clock.  For pesticide/Aroclor analyses performed by
  GC/EC,  the twelve hour time period in  the analytical sequence begins at the
  moment  of injection of the instrument blank that precedes sample analyses, and
  ends  after twelve hours have elapsed according to the system clock.

  VALIDATED TIME OF  SAMPLE RECEIPT  (VTSR)  - the date on which a sample is
  received at  the Contractor's facility, as recorded on the shipper's delivery
  receipt  and  Sample Traffic Report.

  VOLATILE COMPOUNDS -  compounds amenable  to.analysis by the purge and trap
  technique.   Used synonymously with purgeable compounds.
                                        G-8                              OLM04.2

-------
                                                Exhibit G -- Glossary of Terms
WIDE BORE CAPILLARY COLUMN - a gas chromatographic column with an internal
diameter (ID) that is greater than or equal to 0.53 mm.  Columns with lesser
diameters are classified as narrow bore capillary columns.
                                      G-9                              OLM04.2

-------
           EXHIBIT H
AGENCY STANDARD IMPLEMENTATION
              H-l                               OLM04.2

-------
                  Exhibit H - Agency Standard Implementation

                               Table  of  Contents


Section                                                                   Page

1.0   FORMAT CHARACTERISTICS   	  3

2.0   RECORD TYPES  	  5

3.0   PRODUCTION RUNS	6

4.0   RECORD SEQUENCE 	  8

5.0  ' FILE/RECORD INTEGRITY 	  9

6.0   DATES AND TIMES	9

7.0   MULTIPLE VOLUME DATA	9

8.0   DELIVERABLE	10

9.0   RECORD LISTING	12
      9.1   Production Run Header Record  (Type 10)	12
      9.2   Chromatography Record  (Type 11)  	 13
      9.3   Sample Header Data Record (Type  20)	14
      9.4   Sample Header Data Record (Type  21)	16
      9.5   Sample Condition Record  (Type 22)  	 18
      9.6   Associated Injection and Counter Record  (Type 23)	 20
      9.7   Sample Cleanup Record  (Type 27)  	 23
      9.8   Results Data Record  (Type 30)  .  .  .  .-	25
      9.9   Auxiliary Data Record  (Type 32)	28
      9.10  Name Record  (Type  33)	29
      9.11  Instrumental Data Readout Record (Type 36)   	30
      9.12  Comment Record  (Type 90)  	33

10.0  DEFINITIONS OF VARIOUS CODES USED IN AGENCY STANDARD RECORDS   .  .  .  . 34
      10.1  Quality Control and Related Codes  (QCC)  in Type 20 Records   .  . 34
      10.2  Codes For Sample Medium  (Matrix, Sources) 	 36
      10.3  List of Sample and Result Qualifiers	36

APPENDIX A -- FORMAT OF RECORDS FOR SPECIFIC USES	38
      Table of Contents	39
                                      H-2                              OLM04.2

-------
                                                        Exhibit H -- Section 1
                                                        Format Characteristics
1.0   FORMAT CHARACTERISTICS

1.1   This constitutes an implementation of the EPA Agency Standard for
      Electronic Data Transmission based upon analytical results and ancillary
      information required by the contract.  All data generated by a single
      analysis are grouped together, and the groups are aggregated to produce
      files that report data from an SDG.  Because this implementation is only
      a subset of the Agency Standard, some fields have been replaced by
      delimiters as place holders for non-CLP data elements.

1.2   This implementation includes detailed specifications for the required
      format of each record.  The position in the record where each field is
      to be contained relevant to other fields is specified, as well as the
      maximum length of the field.  Each field's required contents are
      specified as literal  (contained in quotes),  which must appear exactly as
      shown  (without quotes), or as a variable for which format and/or
      descriptions are listed in the format/contents column.  Options and
      examples are listed for most fields.  For fields where more than three
      options are available, a list and description of options are supplied on
      a separate page following the record descriptions.  Fields are separated
      from each other by the delimiter "|"  (ASCII 124) .  Fields that do not
      contain data should be zero length or a blank field  (empty with no space
      or additional delimiters between the delimiters before and after the
      field) with the delimiter as a place holder.  For the purposes of
      Section 9 of this exhibit, wherever  "blank" is given as an option under
      the "Format/Contents" column, it refers to a blank field as explained
      above.

1.3   Numeric fields may contain numeric digits, a decimal place, and a
      leading minus sign.  A positive sign is assumed  if no negative sign is
      entered in a numeric  field and shall not be entered into any numeric
      field.  Values that exceed the maximum length allowed shall be reported
      to the maximum possible, maintaining the specified decimal place and
      maximum field length  restrictions.

1.4   Requirements for significant  figures and number  of decimal places are
      specified in Exhibit  B.  The numeric field lengths are specified such
      that  all possible numeric values can be written  to the file.  The size
      of the numeric field  indicates the maximum number of digits, including a
      decimal place and negative sign  (if appropriate), that can appear in the
      field at the same time.  Therefore, the number reported may need.to be
      rounded  (using rounding rules described in Exhibit B) to fit into the
      field.  The rounding  shall maintain  the greatest  significance possible
      providing the field length limitation.  In addition, the rounded number
      that  appears on  the form, and therefore in the field on the diskette
      file,  must be used in any calculation that may result in other numbers
      reported on the  same  form or  other forms  in the  SDG.  The numbers/values
      reported by the  Contractor are used  by CCS to calculate a result  (e.g.,
      CRQL).  The final value calculated by CCS is then rounded according to
      rounding rules described in Exhibit  B and is used for comparison to the
      final value  (e.g., CRQL) reported  by the  Contractor.  Field lengths
      should only be as long as necessary  to contain the data; packing with
      blanks is not allowed.
                                      H-3                               OLM04.2

-------
Exhibit H -- Section 1
Format Characteristics
1.5   The CLP is currently developing a data delivery strategy that may be
      used as an alternative to the requirements stated in Exhibit H.  This
      strategy's intent is to provide a neutral data delivery structure to the
      Contractor that will further facilitate the exchange of analytical
      information, generated under this analytical protocol.  The proposed
      strategy is intended to accommodate laboratories that generate data
      transmission files under multiple data formats.  Upon implementation of
      this alternate electronic data delivery strategy by the CLP and prior to
      submission of data in alternate format(s), the Contractor must first
      demonstrate its ability to provide electronic data as stated in this
      Exhibit H and obtain written permission from the CLP for the submission
      of data in alternate format(s).  The Contractor will receive a written
      response to its request within 90 calendar days. However, until the
      implementation of this alternate electronic data delivery strategy by
      the CLP, all electronic data deliverables must be provided as specified
      in this Exhibit H.
                                      H-4                               OLM04.2

-------
                                                        Exhibit H -- Section 2
                                                                  Record Types
2.0   RECORD TYPES

2.1   The Agency Standard consists of variable length ASCII records.  Maximum
      field length specifications match the reporting requirements in Exhibit
      B.  The last two bytes of each record shall contain "carriage return"
      and "line feed", respectively.

2.2   This implementation consists of twelve record types that can be
      summarized in four groups, designated by the first record type in each
      group:

       Type              Type ID   Contents

       Run Header           10     Information pertinent to a group of
                                   samples processed in a continuous
                                   sequence; usually several per SDG
       Sample Header        20     Sample identifying, qualifying, and
                                   linking information
       Results Record       30     Analyte results and qualifications
       Comments Record      90     Free form comments

2.3   A separate run header is used for volatiles  (VGA), semivolatiles  (SV),
      and for each column analysis for pesticides  (PEST)  (minimum of four type
      10 series for VOA/SV/PEST SDG).  The 20 series records contain sample
      characteristics and link samples within an SDG to the corresponding
      calibrations, blanks, and other QCs.  The 30 series records contain the
      actual analytical results by analyte within each sample.  The 10, 20,
      and 30 records are associated with each other by their position in the
      file  (i.e., 30 series records follow the corresponding 20 series, which
      in turn follow the 10 series run header records).
                                      H-5                               OLM04.2

-------
Exhibit K -- Section 3
Production Runs
3.0   PRODUCTION RUNS

3.1   A production run represents a "group" or "batch" of samples that are
      processed in a continuous sequence under relatively stable conditions.
      Specifically:

3.1.1    Calibration - All samples in a run use the same initial calibration
         data.

3.1.2    Method number - Constant throughout a run.

3.1.3    Instrument conditions - Constant throughout a run.

3.2   Each instrumental analysis consists of a separate production run and is
      reported in a separate file.  There will be a separate production run
      for each of the two pesticide GC columns utilized.  Thus, a full three
      fraction analysis will consist of a minimum of four production runs.
                                     ' Hr6                              OLM04.2

-------
                                                        Exhibit H --  Section 3
                                                               Production Runs
3.3   Example of the Sequence of Record Types in a File1
   10              Contains Run Header information.
       11          Contains additional run-wide information.
       20          Occurs once for each sample, calibration,  mean response
                   factor, matrix spike duplicate result,  etc.   Acts as a
                   header.
       21
       22          Contains additional information for samples.
       23
       27
           30      Occurs once for each final analytical result.  Reports
                   the value being determined as defined by the type 20.
               32  Reports any auxiliary data necessary.
               33  Reports compound names for tentatively identified
                   compounds  (TICs) if necessary.
               36  Reports any instrumental data necessary.
           30      Values for the next analyte or parameter being measured.
               32  Additional data may vary for each parameter,  and may
               33  occur  in any order.  Multiple occurrences of the same
               36  record type, however, must be consecutive.
           30      Continues  for as many as are necessary.
               32
               33
               36
           30
               32
               33
               36
       20          Next Sample Header record.  The following applies to the
       21          next sample or other group of data.
       22
           30
               32
               33
               36
           30
               32
               33
               36
                   etc.
       20
       21
           30
               32
               33                        •
               36
                   etc.
      Appendix A provides  a detailed  set of examples  for the use of the
      different  record types,  and  their relationship  to other record types.

                                      H-7                              OLM04.2

-------
Exhibit H -- Section 4
Record Sequence
4.0   RECORD SEQUENCE

4.1   The sequence of records for Agency Standard files is as follows:  A Run
      Header (type 10) record shall be present once and once only  (per file)
      as the first record in a file.  Therefore, a complete VOA/SV/PEST SDG
      will consist of several files.

4.2   Each environmental sample, calibration standard, or quality control
      sample is represented by a group composed of type 20, 21, 22, 23, and 27
      records,  that hold sample level identifying information, followed by
      type 30,  32, 33, and 36 records for each method analyte including
      surrogates, system monitoring compounds, and internal standards in the
      sample.  The type 20 record holds a count for the number of method
      analytes being determined and includes all target compounds, surrogates,
      system monitoring compounds, and internal standards plus each peak of
      the multi-component pesticides  (do not include TICs in this count).  A'
      separate field on the type 23 record contains the number of TICs found.
      Type 20 records shall occur in the order of sample analysis.  In
      addition, a type 20 record with a QC code "MNC", followed by a type 30
      record for each method analyte  (reporting values such as mean response
      factors)  will appear after the type 10 or type 11 record and before the
      type 20 record that initiates the analytical sequence.  Similarly, for
      pesticide runs, a type 20 record with a QC code "GPC" for GPC recovery,
      followed by type 30 records for each of the method analytes spiked; and
      a type 20 record with a QC code "FLO" for Florisil recovery, followed by
      type 30 records for each of the method analytes (and the two surrogates)
      included in the Florisil check will appear before the type 20 record
      that initiates the analytical sequence.

4.3   Type 90 comment records may be defined to occupy any position after the
      type 10  (header) record.
                                      H-8                              OLM04.2

-------
                                                   Exhibit H -- Sections- 5-7
                    File/Record Integrity/Dates and Times/Multiple Volume Data
5.0   FILE/RECORD INTEGRITY

      All record types shall contain the following check fields to ensure file
      and record integrity:

      Record     Field   Field
      Position   Length  Contents                 Remarks

      First Field    2    Record type              "10" or as appropriate
      Last Field     5    Record sequence number   00001-99999, numbered
                                                 within file sequentially
                     4    Record checksum1         Four hexadecimal digits
                     2    Must contain CR and LF
6.0   DATES AND TIMES

      Date or time-of-day information consists of successive groups of two
      decimal digits  (except year, which is four decimal digits), each
      separated by delimiters.  Dates are given in the order 1TYY MM DD, and
      times as HH MM.  All hours shall be given as 00 to 23 using a 24-hour
      clock and shall be local time.  All days shall be given as 01 to 31.
      All months shall be given as 01 to 12  (e.g., 01 is January, 02 is
      February).

7.0   MULTIPLE VOLUME DATA                             .   '

      There is no requirement under this format that all the data from an
      entire sample delivery group fit onto a single diskette.  However, each
      single production run must fit onto a single diskette if possible.  If
      that is not possible, then it is necessary that all files start with a
      type 10 record, and that the multiple type 10 records for each file of
      the same production run be identical.  Information for a single sample
      shall not be split between files.
       The  checksum is the sum of  the  ASCII  representation of the  data on the
       record up to the Record Sequence  Number (not  including the  Record
       Sequence Number)  plus the checksum of the  previous record.   The sum is
       taken modulo 65536 (216) and is  represented as four hexadecimal digits
       (i.e.,  the remainder of the sum divided by 65536  represented as four
       hexadecimal digits).

                                      H-9                               OLM04.2

-------
Exhibit H -- Section 8
Deliverable
8.0   DELIVERABLE

8.1   The file shall be submitted on IBM-compatible, 3.5 inch high density
      1.44 M-byte diskettes.  The diskettes shall be formatted and recorded
      using MS-DOS Operating System.  The diskettes shall contain all
      information relevant to one and only one SDG.  An alternative means of
      electronic transmission may be utilized if approved in advance by the
      EPA.

8.2   Agency Standard data from an entire SDG may not fit onto a single
      diskette.  If a single production run is being split onto multiple
      diskettes, then all files shall start with a type 10 record, and the
      multiple type 10 records for each file of the same production run shall
      be identical.  Do not split the data from a .single sample.onto multiple
      diskettes.

8.3   Information on the diskette must correspond to information submitted in
      the hardcopy raw data package and on the hardcopy raw data package
      forms.  For example, type 30 results field specifies maximum length of
      13.  When reporting CRQLs or results on Form 1, maximum length is 13 as
      is specified in this exhibit; when reporting  'calculated amounts' on
      Form 7D, hardcopy specified maximum length is 8.  Unused records shall
      not be included on the diskettes.  If the information submitted in the
      hardcopy data package forms is changed, the information in the
      electronic file  (e.g., diskette) shall be changed accordingly, and a
      complete electronic deliverable containing all the information for the
      SDG shall be resubmitted along with the hardcopy at no additional cost
      to the EPA.

8.4   Each diskette shall be identified with an external label containing (in
      this order) the following information:

      Disk Density
      File Name(s)
      Laboratory Name  (optional)
      Laboratory Code
      Contract Number
      Case Number/SDG
      SAS Number  (where applicable)
      Initial Submission or Resubmission  (as applicable) and Date

8.5   The format for File Name shall be XXXXX.O01 to XXXXX.O99.  Where XXXXX
      is the SDG identifier, O designates Organics, and 01 through 99 is .the
      file number.

8.6   Dimensions of the label must be in the range of 2-1/2" to 2-3/4" long by
      2" to 2-1/8" wide for a 3-1/2 inch IBM-compatible diskette.

8.7   Section 9.0  (Record Listing) provides information for the usage of each
      of the record types.  Where specified, labels indicate the nature of the
      value(s) that follow on that record.  If the value(s) will not be
      reported, the label shall be omitted.
                                     H-10                              OLM04.2

-------
                                                        Exhibit H -- Section 8
                                                                   Deliverable

8.7.1    A record type 30 for each TCL compound,  surrogate,  system monitoring
         compound,  and internal standard quantitated for shall be reported.
         If the^TCL is not detected,  the 'U'  qualifier in the appropriate
         field shall be indicative of that.

8.7.2    For multicomponent analytes (Aroclors/toxaphene),  if the
         multicomponent analyte is detected,  a record type 30 and 32 shall be
         reported for each peak identified.
                                      H.-ll                               OLM04.2

-------
Exhibit H -- Section 9
Record Listing
9.0   RECORD LISTING

      The following lists every record type required to report data from a
      single SDG.

9.1   Production Run Header Record  (Type 10)

      Use:      Each production run will start with a record type 10.

       MAXIMUM
       LENGTH    CONTENTS                      FORMAT/CONTENTS
         2       RECORD. TYPE                   "10"
         6       Delimiters                    I  I I I I I
         5       INSTRUMENT/DETECTOR           Character1
         1       Delimiter                     |
         8       METHOD NUMBER                 Character2
         2       Delimiters                    ||
         6       LAB CODE                      Character
         4       Delimiters                    ||||
        11       CONTRACT NUMBER               Character
         1       Delimiter                     |
        10       INSTRUMENT ID                 Character
         2      • Delimiters                    |  |
        25       LABORATORY NAME               Character
         2       Delimiters           .         |  |  '
         5       RECORD SEQUENCE NUMBER        Numeric
         4       CHECKSUM                      Character
      General descriptor  (GC/MS  for VOA/SVOA analysis or GC  for pesticide
      analysis on GC/EC).

      OLM04.2V For Volatiles;  OLM04.2B  for  semivolatiles;  OLM04.2P for
      pesticides.   (O  for Organic, L  for Low, M  for Medium,  zero  four for
      document number,  zero V  for volatiles, zero B for semivolatiles,  zero P
      for pesticides.)

                                      H-12                               OLM04.2

-------
                                                        Exhibit H -- Section 9
                                                                Record Listing
9.2   Chroraatography Record  (Type 11)

      Use:      To describe chromatograph condition.  Must be present once for
                each production run immediately following the record type 10.
                 CONTENTS                      FORMAT/CONTENTS
                 RECORD TYPE                   "11"
                 Delimiter                     |
                 GC COLUMN IDENTIFICATION      Character
                 Delimiters                    ||
                 GC COLUMN ID1                 Numeric  (mm)
                 Delimiters                    I  I I I I I I I 1 I I
                 RECORD'SEQUENCE NO.           Numeric
                 CHECKSUM                      Character
       Internal Diameter of the GC column used.

                                      H-13                              OLM04.2

-------
Exhibit H -- Section 9
Record Listing
9.3   Sample Header Data Record  (Type 20)
          1
          1
          1
          3
          1
          3
          1
          5
          1
          6
          1
          4
          1
          2
          1
          2
          1
          2
          1
          2
          2
          2
          1
          5
           CONTENTS
           RECORD TYPE
           Delimiters
           EPA SAMPLE NUMBER

           Delimiter
           MATRIX
           Delimiter
           QC CODE
           Delimiter
           SAMPLE QUALIFIER
           Delimiter
           CASE NUMBER
           Delimiter
           SDG NO.
           Delimiter
           SAMPLE/BLANK/STANDARDS YEAR ANALYZED
           Delimiter
           SAMPLE/BLANK/STANDARDS MONTH ANALYZED
           Delimiter
           SAMPLE/BLANK/STANDARDS DAY ANALYZED
           Delimiter
           SAMPLE/BLANK/STANDARDS HOUR ANALYZED
           Delimiter
           SAMPLE/BLANK/STANDARDS MINUTE ANALYZED
           Delimiters
           SAMPLE WT/VOL UNITS
           Delimiter
           SAMPLE WT/VOL
FORMAT/CONTENTS
"20"
II
As is exactly on the
hardcopy form
I
CHARACTER1
I
Character (See Section 10)
I
RIN/REX/REJ/SRN/blank2
I
Numeric
I
Character
I
iTYY
I
MM
I
DD
I
HH
I
MM
II
"G"/"ML"/blank3
I
Numeric4
"0" if not applicable (calibration,
soil.
                                           tune,  etc.);  "1"  for water;  "H"  for
       "RIN"  for  reinjection,-  "REX"  for  re -extract ions;  "REJ"  for  rejected
       samples; "SRN"  for dilutions;  and leave blank  (empty field  with  zero
       length) when none of  the  previous conditions apply.   In case  of  multiple
       operations on a sample, the  final operation will  be  indicated (e.g.,
       reinjection of  a dilution; AAA12DLRE would have a QC Code of  "RIN") .

       Sample WT/VOL unit is mL  (milliliters) for liquids and  G (grams)  for
       solids.  The sample units code indicates which units are in use  for the
       current sample.  Leave  blank (zero length) if  not applicable.

       Sample WT/VOL is the  volume  in milliliters for liquid or the  wet weight
       in grams for solids.  Sample WT/VOL includes the  purge  volume. .
                                     H-14
                                                                       OLM04.2

-------
                                                 Exhibit H  -- Section  9
                                                         Record Listing
Sample Header Data Record  (Type 20)  (Cont.)
           CONTENTS                                 FORMAT/CONTENTS
           Delimiter                               |
           ANALYTE COUNT                            Numeric5
           Delimiters                           .    || |
           RECORD SEQUENCE NO.                      Numeric
           CHECKSUM                                 Character
1-3 decimal digits.  Counts TCL analytes, surrogates, system monitoring
compounds  (SMC), internal standards, and all peaks reported for multi-
component PCBs.  Do not include the count for TICs in this field.  For
calibrations, also count DFTPP, if included in calibration solution.

                               H-15                              OLM04.2

-------
  Exhibit H -- Section 9
  Record Listing
  9.4   Sample Header Data Record (Type 21)

        Use:       Continuation of Type 20.
        Position: Follows the Type 20 to which it applies.
MAXIMUM
LENGTH
2
1
1
1
1
2
1
1
1
2
6
1
14
1
4
1'
2
1
•2
2
4
1
2
1
2
2

CONTENTS
RECORD TYPE
Delimiter
PURGE
Delimiter
LEVEL
Delimiters
EXTRACTION
Delimiters
SAS NUMBER
Delimiter
LAB FILE/ SAMPLE ID
Delimiter
YEAR EXTRACTED
Delimiter
MONTH EXTRACTED
Delimiter
DAY EXTRACTED
Delimiters
YEAR RECEIVED
Delimiter
MONTH RECEIVED
Delimiter
DAY RECEIVED
Delimiters
I  2
                                             FORMAT/CONTENTS
                                             "21"
                                             •I
                                             "N" for not heated; "Y" for heated;
                                             blank if SV or PEST
                                             I
                                             "L"/"M"/blank1
                                             II
                                             S/C/H/N/X/P/T/blank (for all other
                                             volatile samples)2
                                             II
                                             Character
                                             I
                                             Character3
                                             I
                                             ITYY/blank  (for volatiles)
                                             I
                                             MM/blank  (for volatiles)
                                             I
                                             DD/blank  (for volatiles)
                                             II
                                             YYYY/blank  (for standards, tunes, and
                                             blanks)
                                             I
                                             MM/blank  (for standards, tunes, and
                                             blanks)
                                             I
                                             DD/blank  (for standards, tunes, and
                                             blanks)
                                             II
"L" for low level samples and "M" for medium level samples for volatile
and semivolatile analyses.  Leave blank for pesticides, all
calibrations, and all tunes.

"S" for separatory funnel;  "C" for continuous liq-liq without
hydrophobia membrane;  "H" for continuous liq-liq with hydrophobia
membrane; "N" for sonication; "X" for automated soxhlet;  "P" for
pressurized fluid; "T" for  volatile low level soils^ by the Modified
Sff-846 Method 5035;.blank (zero length field) for all other volatile
samples.

Lab File ID for volatile and semivolatile analyses.  Lab Sample ID for
pesticides in same format as on:-forms.
                                       H-16
                                                                         OLM04.2

-------
                                                 Exhibit H  -- Section  9
                                                         Record Listing
Sample Header Data Record  (Type  21)  (Cont.)

MAXIMUM
LENGTH    CONTENTS                   FORMAT/CONTENTS
   8      INJECTION/ALIQUOT VOLUME   Numeric/blank  (for  low  level VOA)4
   2      Delimiters       .          | |
   5      RECORD SEQUENCE NO.        Numeric
   4      CHECKSUM                   Character
Injection volume, in uL, for SVOAs and PESTs; Soil Aliquot Volume for
medium level VOA.

                               H-17                              OLM04.2

-------
Exhibit H -- Section 9
Record Listing
9.5   Sample Condition Record  (Type 22)

      Use:      Continuation of type 20. Used to describe  additional  Sample
                Conditions.
      Position: Follows the type 20'and  21 to which  it applies.
MAXIMUM
LENGTH
2
1
1 ^
1
2
1
2
\
2
1
2
1
14
1
4
1
5
1
1
1
8
1
8
3

CONTENTS
RECORD TYPE
Delimiter
CALIBRATION YEAR
Delimiter
CALIBRATION MONTH
Delimiter
CALIBRATION DAY
Delimiter
CALIBRATION HOUR
Delimiter
CALIBRATION MINUTE
Delimiter
CALIBRATION FILE ID
Delimiter
PH
Delimiter
PERCENT MOISTURE
Delimiter
DECANTED
Delimiter
EXTRACT VOLUME
Delimiter
DILUTION FACTOR
Delimiters
                                            FORMAT/CONTENTS
                                            "22"
                                            I
                                            yyYY/blank  (for  PEST)1
                                            I
                                            MM/blank  (for  PEST)
                                            I
                                            DD/blank  (for  PEST)
                                            I
                                            HH/blank  (for  PEST)
                                            I
                                            MM/blank  (for  PEST)
                                            I
                                            Character/blank  (for PEST)2
                                            I
                                            Numeric/blank  (for aqueous  samples
                                            and volatiles)
                                            I
                                            Numeric
                                            I
                                            "Y"/"N"/blank  (for volatiles)
                                            I
                                            Numeric/blank  (for low level VOA)3
                                            I
                                            Numeric4
                                            III
      For volatiles and  semivolatiles,  enter  the  date  and time  of analysis  of
      the most recent  50 ug/L  (VOAs)  or the 50  ng (SVGAs)  standard run prior
      to the sample reported in the  associated  type  20 record.   Leave  blank
      for pesticides.

      Lab File ID of standard  specified in 1  above  (volatiles/semivolatiles
      only).  This field must  match  the Lab File  ID  on Type  21  for the
      associated calibration  (VSTD050/SSTD050).   Leave blank for pesticides.

      Enter the Soil Extract Volume  for medium  level VOA,  and Concentrated
      Extract Volume for all SVOA and PEST.   The  value should be reported in
      microliters.

      Dilution factor  of sample analyzed (omit  contract-mandated dilutions).

                                      •H-18                               OLM04.2

-------
                                                 Exhibit H  -- Section  9
                                                         Record Listing
Sample Condition Record  (Type 22)  (Cont.)
          'CONTENTS                   FORMAT/CONTENTS
          LEVEL                      Numeric/blank  (for VOA/SV)5
          Delimiter                  |
          RECORD SEQUENCE NO.        Numeric
          CHECKSUM                   Character
Concentration level of Pesticide Individual Mix A and B standards.
Concentration of low point, mid point and high point calibration
standards as a multiplier of low point.  Low point = 1.0; Mid point =
4.0; High point 2 16.0.

                               H-19                              OLM04.2

-------
Exhibit H -- Section 9
Record Listing


9.6   Associated Injection and Counter Record (Type 23)

      Use:      Continuation of type 20. Used to identify associated blanks
                and tunes, and the number of surrogates/SMCs and spikes
                outside of the QC limits and the number of TICs.
      Position: Follows the type 20, 21, and 22 to which it applies.

       MAXIMUM
       LENGTH    CONTENTS                 FORMAT/CONTENTS
          2      RECORD TYPE              "23"
          1      Delimiter                |
          1      INSTRUMENT PERFORMANCE   "P" (for BFB and DFTPP IPC) or blank (for
                 CHECK  (IPC/TUNE) LABEL   pesticides)
          1      Delimiter                |
    |      4      IPC/TUNE INJECTION YEAR  YYYY/blank  (for PEST)
          1      Delimiter                |
          2      IPC/TUNE INJECTION       MM/blank  (for PEST)
                 MONTH
          1      Delimiter                |
          2      IPC/TUNE INJECTION DAY   DD/blank  (for PEST)
          1      Delimiter                |
          2      IPC/TUNE INJECTION HOUR  HH/blank  (for PEST)
          1      Delimiter                |
          2      IPC/TUNE INJECTION       MM/blank  (for PEST)
                 MINUTE
          1      Delimiter                |
         14      DFTPP/BFB LAB FILE ID    Character/blank  (for  PEST)
          1      Delimiter                |
          2      VOLATILE STORAGE BLANK   "HB"  (for VOA) or blank  (for SV and PEST)
                 LABEL
          1      Delimiter                |
    |      4      STORAGE BLANK INJECTION  ITYY/blank  (for SV and PEST)
                 YEAR
          1      Delimiter                |
        '  2      STORAGE BLANK INJECTION  MM/blank  (for SV and  PEST)
                 MONTH
          1      Delimiter                |
          2      STORAGE BLANK- INJECTION  DD/blank  (for SV and  PEST)
                 DAY
          1      Delimiter                |
          2      STORAGE BLANK INJECTION  HH/blank  (for SV and  PEST)
                 HOUR
          1      Delimiter                |
          2      STORAGE BLANK INJECTION  MM/blank  (for SV and  PEST)
                 MINUTE
          1      Delimiter                |
         14      STORAGE BLANK LAB        Character
                .FILE ID  (VOA ONLY)
          4     . Delimiters
                                     H-20                              OLM04.2

-------
                                                 Exhibit H  --  Section 9
                                                         Record Listing
Associated Injection and Counter Record  (Type 23)  (Cont.)
MAXIMUM
LENGTH
2
1
1 ^

1
2

1
2

1
2

1
2

1
14



1
1


1
2


1
1
1
2
1
1
1

CONTENTS
METHOD BLANK LABEL
Delimiter
METHOD BLANK INJECTION
YEAR
Delimiter
METHOD BLANK INJECTION
MONTH
Delimiter
METHOD BLANK INJECTION
DAY
Delimiter
METHOD BLANK INJECTION
HOUR
Delimiter
METHOD BLANK INJECTION
MINUTES
Delimiter
METHOD BLANK LAB
FILE (for VOA and
SV) /SAMPLE ID (for
PEST)
Delimiter
SURROGATE (for SV and
PEST) /SMC (for VOA)
RECOVERY LABEL
Delimiter
SURROGATE (for SV and
PEST) /SMC (for VOA)
RECOVERIES OUT
Delimiter
TIC LABEL
Delimiter
NO. OF TICS
Delimiter
SPIKE RECOVERY LABEL
Delimiter
                                    FORMAT/CONTENTS
                                    "MB"/blank  (for  standard,  tune,  and
                                    method blanks)
                                    I
                                    YYYY/blank  (for  standard,  tune,  and
                                    method blanks)
                                    I
                                    MM/blank  (for standard,  tune,  and method
                                    blanks)
                                    I
                                    DD/blank  (for standard,  tune,  and method
                                    blanks)

                                    HH/blank  (for standard,  tune,  and method
                                    blanks)
                                    I
                                    MM/blank  (for standard,  tune,  and method
                                    blanks)
                                    I
                                    CHARACTER
                                    "P" for % recoveries/blank  (for STD/IPC)
                                    Numeric1
                                    "T" (for VOA and SV TICs)/blank  (for
                                    PEST)
                                    I
                                    Numeric
                                    I
                                    "S" for Matrix Spikes and Matrix Spike
                                    Duplicates/blank for anything else
This will be the number of surrogate  (for SV or PEST) or SMC  (for VOA)
recoveries outside QC limits for a specific column.  It should not be
cumulative of the two columns for pesticides.
                               H-21
                                                                 OLM04.2

-------
Exhibit H -- Section 9
Record Listing
       Associated Injection and Counter Record  (Type 23)  (Cont.)

       MAXIMUM
       LENGTH    CONTENTS                 FORMAT/CONTENTS
          2      SPIKE RECOVERIES OUT     Numeric/blank2
          1      Delimiter                |
          1      RPD LABEL                "R" for RPD/blank3
          1      Delimiter                |
          2      RPD OUT                  Numeric
          1      Delimiter                |
          5      RECORD SEQUENCE 'NO.      Numeric
          4      CHECKSUM                 Character
      Enter the number of  spike recoveries out.  Enter  "0"(zero) if none of
      the spike recoveries  are outside of the QC limit.

      "R" for Matrix Spike/Matrix Spike Duplicate Recovery Relative Percent
      Differences.  Leave  blank for all other samples  (only report for
      MS/MSD).

                                     H-22                              OLM04.2

-------
                                                        Exhibit H -- Section 9
                                                                Record Listing
9.7   Sample Cleanup Record (Type 27)

      Use:      Continuation of type 20. Used to identify sample/blank cleanup
                procedures and QC results.
      Position: Follows type 20, 21, 22, and 23 to which it applies.
                                                   FORMAT/CONTENTS
                                                   "27"
                                                   I
                                                   "G" for GPC/blank  (for VGA}1
                                                   I
                                                   ITYY/blank  (for VGA)
                                                   I
                                                   MM/blank  (for VGA)
                                                   I
                                                   DD/blank  (for VGA)
                                                   I
                                                   HH/blank  (for VGA)
                                                   I
                                                   MM/blank  (for VGA)
                                                    I
                                                   Character/blank  (for VGA and
                                                   SV)2
                                                    I
                                                   "F"  (for  PEST) or  blank  (for
                                                   VGA and SV)
                                                    I
                                                   ITYY/blank  (for VGA and  SV)
                                                    I
                                                   MM/blank  (for VGA  and SV)
                                                    I
                                                   DD/blank  (for VGA  and SV)
                                                    I
                                                   HH/blank  (for VGA  and SV)
                                                    I
                                                   MM/blank  (for VGA  and SV)
                                                    I
                                                   Character3
       "G"  indicates  that GPC was performed.   If GPC was  not  performed,.leave
       the  field blank.

       Lab  Sample ID  of  associated GPC.   This is a unique identifier assigned
       to the spike recovery results for a specific GPC calibration check for
       pesticides.  Leave blank for volatiles and semivolatiles.

       Lab  Sample ID  of  associate Florisil lot check.   This  is a  unique
       identifier assigned to a lot of Florisil cartridges.
MAXIMUM
LENGTH
2
1
1
1
1 «
1
2
1
2
1
2
1
2
1
14
1
1
1
1 *
1
2
1
2
1
2
1
2
1
14

CONTENTS
RECORD TYPE
Delimiter
FIRST CLEANUP TYPE
Delimiter
GPC CALIBRATION CHECK YEAR
Delimiter
GPC CALIBRATION CHECK MONTH
Delimiter
GPC CALIBRATION CHECK DAY
Delimiter
GPC CALIBRATION CHECK HOUR
Delimiter
GPC CALIBRATION CHECK MINUTE
Delimiter
GPC Data Descriptor
Delimiter
.FLORISIL CLEANUP TYPE
Delimiter
FLORISIL LOT CHECK YEAR
Delimiter
FLORISIL LOT CHECK MONTH
Delimiter
FLORISIL LOT CHECK DAY
Delimiter
FLORISIL LOT CHECK HOUR
Delimiter
FLORISIL LOT CHECK MINUTE
Delimiter
FLORISIL DATA DESCRIPTOR
                                      H-23
                                                                        OLM04.2

-------
Exhibit H -- Section 9
Record Listing
       Sample Cleanup Record  (Type 27)  (Cont.)

       MAXIMUM
       LENGTH  .  CONTENTS
          1      Delimiter
          1      SULFUR CLEANUP •

          1      Delimiter
          2      SULFUR BLANK LABEL
          1      Delimiter
          4      SULFUR BLANK INJECTION YEAR
          1      Delimiter
          2      SULFUR BLANK INJECTION MONTH
          2      Delimiters
          2      SULFUR BLANK INJECTION DAY
          1      Delimiter
          2      SULFUR BLANK INJECTION HOUR
          1      Delimiter
          2      SULFUR BLANK INJECTION MINUTE
          1      Delimiter
         14      SULFUR BLANK LABORATORY/
                 SAMPLE ID
          1      Delimiter
          5      RECORD SEQUENCE NO.
          4      CHECKSUM
FORMAT/CONTENTS
I
Y/N (for PEST)/blank (for VOA
and SV)
11 SB"/blank  (if no separate
sulfur blank was prepared for
pesticides; also blank for VOA
and SV)
I
iTYY/blank  (for VOA and SV)
I
MM/blank  (for VOA and SV)
II
DD/blank  (for VOA and SV)
I
HH/blank  (for VOA and SV)
I
MM/blank  (for VOA and SV)
I
Character

I
Numeric
Character
                                     H-24
                                                                       OLM04.2

-------
9.8 .  Results Data Record (Type 30)
                                                        Exhibit H -- Section 9
                                                                Record Listing
                 CONTENTS
                 RECORD TYPE
                 Delimiter
                 ANALYTE LABEL
          1      Delimiter
          9      CAS NUMBER
          1      Delimiter
          9      INTERNAL STD. CAS NUMBER
          1      Delimiter
          5      CONCENTRATION UNITS
          1      Delimiter
          3      RESULT QUALIFIER
          1      Delimiter
         13      RESULTS
          1      Delimiter
          5      FLAGS
          1      Delimiter
          1      AMOUNT ADDED  LABEL
          1      Delimiter
         13      AMOUNT ADDED
          1      -Delimiter
FORMAT/CONTENTS  '
"30"
I
"C" for CAS Number  (blank for
unknown TICs)
I
Numeric (for TCL, surrogates,
DFTPP, BFB, SMC, internal.
standards, and identified TICs)

Numeric

Character  "ug/L" (aqueous);
"ug/Kg" (soil) ,-  "ng"  (amount
added)

Character1'2
I
Numeric3
I
Character4
I
"A" for Amt. added5
I
Numeric
1     When  a  Type  20  Record is  used  for calibration  summary (MNC),  the
      associated Type 30  Record uses "AVG"  for  average  RRFs and Mean
      Calibration  Factors.   See Exhibit H Section  10.3.2.

2     For pesticide sample  analysis,  if an analyte is detected in only  one  of
      the two column  analyses,  report the analyte  as "not  detected." in  both
      runs.   Report result  qualifier,  for each  column,  as  BDL.  See Section
      10.3.2  for result qualifiers.

3     Leave this field blank only when reporting non-detects.

4     A maximum of five flags (D,E,J,B,A,P,C,X,Y,Z,  or  N)  with no space
      between the  flags can be  reported,  each representing a qualification  of
      the result as described in Exhibit B.  For surrogates, the "D" flag will
      indicate surrogates diluted out.

5     For Matrix Spike/Matrix Spike  Duplicate analysis,  surrogate,  SMC  for
      VOA,  SV,  and PEST  (Form 3s).   Nominal Amount for  Pesticides (Form
       7E/7F) .  Spike  added  for  florisil and GPC (Form 9A/9B) .
                                      H-25
                         OLM04.2

-------
Exhibit H -- Section 9
Record Listing
Results Data Record  (Type 30)  (Cont.)

MAXIMUM
LENGTH    CONTENTS
1
I
13
1
1
1
5
1
1
1
5
1
5
1
1
1
5
1
1
1
5
CRQL LABEL
Delimiter
CRQL
Delimiter
RSD LABEL
Delimiter
RSD VALUE
Delimiter
MS/MSD REC LABEL
Delimiter
MS % RECOVERY
Delimiter
MSD % RECOVERY
Delimiter
RPD LABEL
Delimiter
RPD VALUE
Delimiter
SURR/ SPIKE RECOVERY LABEL
Delimiter
SURR/SPIKE RECOVERY
                                                FORMAT/CONTENTS
                                                "U"  for "undetected"  or blank when
                                                analyte is detected
                                                I
                                                Numeric
                                                "R"  for %  Resolution/RSD6
                                                I
                                                Numeric
                                                I
                                                "P"  for %  recovery [MS/MSD] /blank
                                                (for sample [except MS/MSD] ,
                                                standard,  tune,  blanks,
                                                calibration)
                                                Numeric/blank (for everything
                                                except  MS)
                                                I
                                                Numeric/blank (for everything
                                                except  MSD)
                                                I
                                                "D"  for MS/MSD or for pesticide
                                                calibration verification
                                                (%D) /blank
                                                I
                                                Numeric/blank7
                                                I
                                                "S"  for % recovery/blank (for non-
                                                surrogate/SMC and non-spike
                                                analytes
                                                I
                                                %  Recovery/blank8
       "R" for  % Resolution  (Forms 6H,  61,  6J,  and 6K)  or for RSD of Response
       factors  under  Calibration  summary (MNC)  Type 20.   (Blank for VOA and SV
       fractions.)

       RPD for  MS/MSD recoveries,  or %D for pesticides.   Calibration
       Verification  (Form  7E/7F).   Otherwise,  leave blank.

       Surrogate  (for SV and PEST)/SMC  (for VOA)  or Spike (Forms 2,  Form 9A/9B)
       recovery.  Leave blank for non-surrogate and non-spike analytes.
                                      H-26
                                                                        OLM04.2

-------
      Results Data Record  (Type 30)  (Cont.)
                                                       Exhibit H  -- Section  9
                                                               Record  Listing
      MAXIMUM
      LENGTH    CONTENTS
          1      Delimiter
FORMAT/CONTENTS
                MEAN  CONCENTRATION  LABEL
          1      Delimiter
         13      MEAN CONCENTRATION

          1      Delimiter
          1      PERCENT DIFFERENCE.LABEL

          1      Delimiter
          5      PERCENT DIFFERENCE
          1      Delimiter
          1      INTERNAL STANDARD AREA LABEL

          1      Delimiter
         13      INTERNAL STANDARD AREA

          1      Delimiter
          5      RECORD SEQUENCE NO.
          4      CHECKSUM
"M" for Mean cone,  (for
multicomponent PEST only)/blank
(for VOA and SV)
I
Numeric (for PEST)/blank (for VOA
and SV)9

"F" or  "P"  (PEST)/blank  (for VOA
and SV  field sample analysis)10
I
Numeric
I
"I" for IS Area  (for VOA and
SV)/blank  (for PEST)
I
Numeric (for VOA and SV)/blank
(for PEST)
I
Numeric
Character
9     Mean Concentration for Multicomponent analytes detected in pesticide
      analyses.

10    "P" for Percent Difference between concentrations from two columns in
      pesticide analyses, or "F" for Percent Difference between average RRF
      (initial, calibration) and RRF50 (continuing calibration) in VOA/SVOA
      analyses.   Leave blank for volatile and semivolatile sample, blank, and
      tune analysis.
                                     H-27
                                                                       OLM04.2

-------
Exhibit H -- Section 9
Record Listing
9.9   Auxiliary Data Record  (Type 32)

      Use:      Used to report retention time  (in minutes) for Internal
                Standards and for TICs  (for Volatiles and Semivolatiles). Used
                to report retention time data and percent breakdown  (for
                pesticides).
      Position: Follows type 30.  (Record will only be required as specified
                above.)
       MAXIMUM
       LENGTH  .  CONTENTS
          2      RECORD TYPE
          3      Delimiters
          2      RETENTION TIME LABEL
          1      Delimiter
          5      RETENTION TIME
          1      Delimiter
          3      FIRST LIMIT LABEL
          1      Delimiter
          5      RT WINDOW LOWER LIMIT
          1      Delimiter
          3      SECOND LIMIT LABEL
          1      Delimiter
          5      RT WINDOW UPPER LIMIT
          2     • Delimiters
          2      % BREAKDOWN LABEL

          1      Delimiter
          5      % BREAKDOWN
FORMAT/CONTENTS
"32"
Ml
"RT"
I
Numeric
I
"RTF"
I
Numeric
I
"RTT"
I
Numeric
II
"PB" for % breakdown/blank (for
VOA and SV)
I
Numeric (DDT/ENDRIN)/blank (for
VOA and SV)
          1      Delimiter
          5      COMBINED % BREAKDOWN
          2      Delimiters
          1      PEAK
          1      Delimiter
          5      RECORD SEQUENCE NO.
          4      CHECKSUM
Numeric/blank (for VOA and SV)1
II
1 THROUGH 5 (for pesticide
multicomponent compounds)/blank
(for VOA and SV)2 .
I
Numeric
Character
      The combined %breakdown will be reported on both the record type 32s
      (for DDT and Endrin).

      For positively identified compounds, a minimum of 3 peaks and a maximum
      of 5 peaks are allowed.  Types 30 and 32 will be repeated for each peak
      that is reported  (a minimum of three, a maximum of five times).  This is
      for multicomponent analytes in pesticide analyses.
                                     H-28
                                                                       OLM04.2

-------
                                                        Exhibit H -- Section 9
                                                                Record Listing
9.10  Name Record (Type 33)

      Use:      This record type is used for volatile and semivolatile
                analyses only to carry an analyte name for TICs.  This record
                is not used for pesticide analysis.
      Position: Follows types 30 and 32 for TICs.
                 CONTENTS                      FORMAT/CONTENTS
                 RECORD TYPE                   "33"
                 Delimiter                     |
                 NAME OF COMPOUND              Character
                 Delimiter                     |
                 RECORD SEQUENCE NO.           Numeric
                 CHECKSUM                      Character
                                      H--29                               OLM04.2

-------
Exhibit H -- Section 9
Record Listing
9.11  Instrumental Data Readout Record  (Type 36}

      Use:      This record type is only used for volatile and semivolatile
                analyses to describe DFTPP/BFB percent abundances.  This
                record is not used for pesticide analysis.
      Position: Follows type 30 for DFTPP/BFB data.
       MAXIMUM
       LENGTH    CONTENTS
          2      RECORD TYPE
          1      Delimiter
          1      MASS LABEL
          3      Delimiters
          3      FIRST MASS  (DFTPP/BFB)
          2      Delimiters
          5      FIRST PERCENT RELATIVE
                 ABUNDANCE
          1      Delimiter
          3      SECOND MASS
          1      Delimiter
          5      SECOND PERCENT RELATIVE
                 ABUNDANCE
          1      Delimiter
          5      PERCENT MASS OF 69
          1      Delimiter
          3      THIRD MASS
          1      Delimiter
          5      THIRD PERCENT RELATIVE
                 ABUNDANCE
          2      Delimiters
          3      FOURTH MASS
          1      Delimiter
          5      FOURTH PERCENT RELATIVE
                 ABUNDANCE
          1      Delimiter
          5      PERCENT MASS OF 69
          1      Delimiter
          3      FIFTH MASS
          1      Delimiter
          5      FIFTH PERCENT RELATIVE
                 ABUNDANCE
          1      Delimiter
          5      PERCENT MASS OF 174
          1      Delimiter
          3      SIXTH MASS
          1      Delimiter
FORMAT/CONTENTS
"36"
I
"M"
III   ,
Numeric (DFTPP for SV or BFB for VGA)
II
Numeri'c

I
Numeric
I
Numeric

I
Numeric, DFTPP only/blank  (for VOA)
I
Numeric
I
Numeric

II  .'
Numeric
I
Numeric

I
Numeric, DFTPP only/blank  (for VOA)

Numeric
I
Numeric

I
Numeric, BFB only/blank (for SV)

Numeric
                                     H--30
                            OLM04.2

-------
                                                 Exhibit H -- Section 9
                                                         Record Listing
Instrumental Data Readout Record (Type 36) (Cont.)
MAXIMUM
LENGTH    CONTENTS
   5      SIXTH PERCENT RELATIVE
          ABUNDANCE
   2      Delimiters
   3      SEVENTH MASS
   1      Delimiter
   5      SEVENTH PERCENT RELATIVE
          ABUNDANCE
   1      Delimiter
   5      PERCENT MASS OF 174
   1      Delimiter
   3      EIGHTH MASS
   1      Delimiter
   5      EIGHTH PERCENT RELATIVE
          ABUNDANCE
   1      Delimiter
   5      PERCENT MASS OF 174
   1      Delimiter
   3      NINTH MASS
   1      Delimiter
   5      NINTH PERCENT RELATIVE
          ABUNDANCE
   1      Delimiter
   5      PERCENT MASS OF 176
   1      Delimiter
   3      TENTH MASS
   1      Delimiter
   5      TENTH PERCENT RELATIVE
          ABUNDANCE
   2      Delimiters
   3      ELEVENTH MASS
   1      Delimiter
   5      ELEVENTH PERCENT
          RELATIVE ABUNDANCE
   2      Delimiters
   3      TWELFTH MASS
   1      Delimiter
   '5      TWELFTH  PERCENT RELATIVE
          ABUNDANCE
   2      Delimiters
   3      THIRTEENTH  MASS
   2      Delimiters
FORMAT/CONTENTS
Numeric
Numeric

Numeric

I                         "
Numeric, BFB only/blank  (for SV)
I
Numeric
I
Numeric

I
Numeric, BFB only/blank  (for SV)
I
Numeric
I
Numeric


Numeric, BFB only/blank  (for SV)
I
Numeric/blank  (for VOA)
I
Numeric/blank  (for VOA)

II
Numeric/blank  (for VOA)

Numeric/blank  (for VOA)

 II
Numeric/blank  (for VOA)
 I
Numeric/blank  (for VOA)
 Numeric/blank (for VOA)
 II
                               H-31
                                                                 OLM04.2

-------
Exhibit H -- Section 9
Record Listing
       Instrumental Data Readout Record  (Type 36)  (Cont.)

       MAXIMUM
       LENGTH    CONTENTS                  FORMAT/CONTENTS
          5      THIRTEENTH PERCENT        Numeric/blank  (for VOA)
                 RELATIVE ABUNDANCE
          1      Delimiter                 |
          5      PERCENT MASS OF 442       Numeric, DFTPP only  (blank for VOA)
          1      Delimiter                 |
          5      RECORD SEQUENCE NO.       Numeric
          4      CHECKSUM                  Character
                                     H-32 .                             OLM04.2

-------
                                                        Exhibit H -- Section 9
                                                                Record Listing
9.12  Comment Record (Type 90)

      Use:      To provide for operator-entered comments.
      Position: May occur anywhere in the file after the type 10 record.

       MAXIMUM
       LENGTH    CONTENTS                      FORMAT/CONTENTS
          2      RECORD TYPE                   "90"
          1      Delimiter                     |
         67      ANY COMMENT                   Character
          1      Delimiter                     |
          5      RECORD SEQUENCE NO.           Numeric
          4      CHECKSUM                      Character
                                     -H-33                               OLM04.-2

-------
Exhibit H -- Section 10
Definitions of Various Codes
10.0  DEFINITIONS OF VARIOUS CODES USED IN AGENCY STANDARD RECORDS

10.1  Quality Control and Related Codes  (QCC) in Type 20 Records

10.1.1   Note:  These codes appear in the QC code fields of type 20 records.
         They are. used to indicate the type of data that is being reported.
         QCC  Name

         LRB  LABORATORY  (REAGENT)
              BLANK

         LIB  LABORATORY  INSTRUMENT
              BLANK

         LSB  LABORATORY  SULFUR BLANK
          LHB   LABORATORY STORAGE  BLANK

          FRB   FIELD BLANK
          FRM  FIELD REFERENCE SAMPLE
Definition

The "Method Blank"  (see Exhibit G).


The "Instrument Blank".


If different from "Method Blank"
(pesticides).

The storage blank (volatiles).

This is any sample  that is submitted
from the field and  is identified as
a blank.  This includes trip blanks,
rinsates, equipment blanks, etc.

This is any sample  that is submitted
for a Case and is identified as a
Performance Evaluation  (PE) sample.

An 'environmental sample which is
analyzed according  to the analytical
method, and subsequently used for
the matrix spike and the matrix
spike duplicate  (see Exhibit G).

The "Matrix Spike"  (see Exhibit G);
must precede LF2.

The "Matrix Spike Duplicate" (see
Exhibit G).
          LSD  LABORATORY SPIKE
               DUPLICATE  BACKGROUND
               (ORIGINAL)  VALUES
          LF1   LABORATORY SPIKED SAMPLE
               -  FINAL -  FIRST MEMBER

          LF2   LABORATORY SPIKED SAMPLE
               -  FINAL -  SECOND MEMBER
          LPC  LABORATORY PERFORMANCE
               CHECK SOLUTION
A solution of DFTPP  (SVGA) or BFB
(VGA) or method analytes  (PEST/PCB)
used to evaluate the performance of
an instrument with respect to a
defined set of criteria  (Tune or
Resolution Check Sample)  (see
Exhibit G).
          FLO  FLORISIL CHECK SOLUTION
A solution of pesticides used to
check recovery from each lot of
Florisil cartridges.  These recovery
results will  be provided in every
production run where associated
samples are analyzed.
                                     H-34
                                                                       OLM04.2

-------
                                                       Exhibit H  --  Section 10
                                                  Definitions of  Various  Codes
         GPC  GPC CHECK SOLUTION
                                A solution of pesticides used to
                                check recovery from each new GPC
                                calibration.  These recovery results
                                will be provided in every production
                                run where associated samples are
         CLM  INITIAL CALIBRATION -
              MULTI-POINT
         CLS  INITIAL CALIBRATION
              SINGLE POINT


         CLC  CONTINUING CALIBRATION
              CHECK

         CLE  CONTINUING PERFORMANCE
              CHECK
10.1.2
                                The Initial Calibration for GC/MS
                                (see Exhibit G),  or the Initial
                                Individual Standard Mixes (A, B) for
                                pesticides (see Exhibit D PEST).
                                Response factors (GC/MS) or
                                Calibration Factors  (pesticides)
                                will be reported on the following
                                type 30 records.

                                The Initial Toxaphene/Aroclor Mixes
                                used to determine all calibration
                                factors  (see Exhibit D PEST).

                                The continuing calibration
                                (VSTDOSO/SSTD050) for GC/MS.

                                The subsequent Individual Standard
                                Mixes  (A,B), Performance Evaluation
                                Mixture, and for subsequent
                                injections of Toxaphene/Aroclor
                                mixes  for pesticides  (see Exhibit D
                                PEST).

                                A calibration solution as above used
                                both as an initial calibration  (CLM)
                                and a  continuing check  (CLC).   (50
                                level  initial calibration if needed
                                for Form 8.)
The following QCC values are used on type 20 records which act as a
header, and indicate that additional (usually calculated) analyte
specific data will be present on type 30  (and following' type)
records.  Usually, these data will apply to an entire production run,
in which case they will appear immediately following the type 10
record or type 11 record if present.  If the data apply to only a
portion of the samples in the run, they shall be placed immediately
preceding the samples to which they apply.  Much of the rest of the
information in the type 20 record may be blank, indicating that these
data do not apply to these results.
         CLD  DUAL PURPOSE CALIBRATION
         MNC   MEAN VALUES FROM
               CALIBRATIONS
                              The data  following represent mean
                              values and percent RSDs from the
                              initial calibration  (GC/MS) or the
                              mean calibration  factors, mean
                              retention times and  retention time
                              windows  (pesticides).
                                     H-35
                                                              OLM04.2

-------
Exhibit H -- Section 10
Definitions of Various Codes
10.2  Codes For Sample Medium  (Matrix, Sources)

            Medium                                       Code
            All Media, Specific Medium not Applicable.   0  (zero)
                 Use for Calibrations, Tunes, etc.
            Water                                        1
            Soil                                         H

10.3  List of Sample and Result Qualifiers

      Definition:  A sample qualifier consists of three characters which act
      as an indicator of the fact and the reason that the subject analysis  (a)
      did not produce a numeric result, or  (b) produced a numeric result for
      an entire sample but it  is qualified in some respect  relating to the
      type or validity of the  result.

10.3.1   Sample Qualifiers


         Oualifier   Full Name    Definition

         RIN         RE-ANALYZED   The indicated analysis results were
                                   generated from a  re-injection of the same
                                   sample extract or aliquot  (RE SUFFIX).

         REX         RE-PREPARED   The indicated analysis results were
                                   generated from a  re-extraction of  the same
                                   sample  (RE SUFFIX).

         REJ         REJECTED      The results for the entire  sample  analysis
                                   have been rejected for an unspecified
                                   reason by the laboratory.   For initial
                                   calibration data,  these  data  were  not
                                   utilized in the calculation of the mean.

         SRN         DILUTED       The indicated analysis results were
                                   generated from a  dilution of  the same
                                   sample  (DL SUFFIX).

10.3.2   Result Qualifiers in  Type 30 Records

         A result qualifier consists of three characters which act as an
         indicator of the fact and the reason that the subject analysis  (a)
         did not produce a numeric result, or  (b) produced  a numeric  result
         for a single analyte  but it is qualified in some respect relating  to
         the type or validity  of the result.  This qualifier is  complementary
         to the flags field on a type 30 record.  A  TIC must have either a
         TIE, TFB, ALC, or PRE result qualifier.
         BDL  ' BELOW DETECTABLE         Indicates  compound was analyzed for but
              LIMITS                   not  detected (Form 1  "U"  Flag).

         NAR  NO ANALYSIS  RESULT      There  is no  analysis  result  required
                                       for  this subject  parameter.

                                     H-36                              OLM04.2

-------
                                                Exhibit H  --  Section  10
                                           Definitions of  Various  Codes
  AVG  AVERAGE VALUE
   CBC  CANNOT BE CALCULATED
   LTL  LESS THAN LOWER
        CALIBRATION LIMIT
   GTL  GREATER THAN UPPER
        CALIBRATION LIMIT
   LLS  LESS THAN LOWER
        STANDARD
   TIE  TENTATIVELY IDENTIFIED
        ESTIMATED VALUE
   REJ  REJECTED


   STD  INTERNAL STANDARD


   STB  INTERNAL STANDARD
        BELOW DETECTION LIMITS

   FBK  FOUND IN BLANK
   TFB  TENTATIVELY IDENTIFIED
|        AND FOUND IN BLANK

   ALC  ALDOL CONDENSATION
   NRP  NON-REPRODUCIBLE
   PRE  PRESUMPTIVE PRESENCE
Average value -- used to report a range
of values  (e.g., relative response
factors).

The analysis result cannot be
calculated because an operand value is
.qualified  (e.g., identifies analytes
whose internal standard is not found)
 (Form 1  "X" Flag).

Analysis result  is from a diluted
sample  (DL suffix) and may be less
accurate than the result from an
undiluted  sample (Form 1 "D" Flag).

Actual value is  known to be greater
than the upper calibration range  (Form
1  "E" Flag).

The analysis result is less than the
sample guantitation limit  (Form 1  "J"
Flag).

The indicated analyte is a tentatively
identified analyte; its concentration
has been'estimated  (Form  1-F.or 1-G
 "J" Flag).

Results  for the  analyte are rejected by
the laboratory.

The indicated compound is an internal
 standard.

A  combination of "STD" and  "BDL".
 The indicated compound was found in the
 associated method blank (LRB)  as well
 as  the sample (Form 1 "B"  Flag).

 A Combination of "TIE" and "FBK"  (Form
 1-F or 1-G "B" Flag).

 Labels a suspected Aldol Condensation-
 product for TICs  (Form 1-G "A"  Flag).

 Results of two or more injections are
 not comparable (Form l-£ "P" flag),
 e.g.,  Aroclor target analyte with
 greater than 25% difference between
 mean concentrations of the two column
 analyses.

 Presumptive evidence of presence of
 material for TIC (Form 1-F or 1-G "N"
 Flag).
                               H-37
                                                                 OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
               APPENDIX A --  FORMAT OF  RECORDS  FOR  SPECIFIC  USES
The USEPA does not warrant or guarantee the completeness and/or accuracy of the
representative examples of  record  type uses  provided in this appendix.    This
appendix  serves  as  an example for  the usage of  record  types and in  no way
redefines or supersedes the specifications or requirements stated in Exhibits A
through H of OLM04.2.  NOTE:  Examples  are representative and are not typically
provided for both columns used in the pesticide analysis.
                                     H-38                              OLM04.2

-------
                                                       Exhibit H -- Appendix A
                                           Format of Records for Specific Uses
               Appendix A --  Format  of  Records  for  Specific Uses

                               Table of Contents

Section                                                               '    Page

1.0   VOLATILES	42
      1.1   Record Type '10 for Volatiles	42
      1.2   Record Type 11 for Volatiles	42
      1.3   Record Type 20s and 30s for Volatiles  --
            Initial Calibration Mean Values  (QC Code  'MNC')	42
      1.4   Record Type 20s and 30s for Volatiles  --
            BFB Tune  (QC Code  'LPC')	42
      1.5   Record Type 20s and 30s for Volatiles  --
            Initial Calibration Standard  (QC Code  'CLM')  	 42
      1.6   Record Type 20s and 30s for Volatiles  --
            Dual Purpose Calibration Standard  (QC  Code  'CLD')  	 43
      1.7   Record Type 20s and 30s for Volatiles  --
            Continuing  Calibration  Standard  (QC Code  'CLC')   	 43
      1.8   Record Type 20s and 30s for Volatiles  --
            Method Blank  (QC Code 'LRB')   	44
      1.9   Record Type 20s and 30s for Volatiles  --
            Storage Blank  (QC  Code  'LHB')	45
      1.10  Record Type 20s and 30s for Volatiles  --
            Instrument  Blank  (QC Code  'LIB')	46
      1.11  Record Type 20s and 30s for Volatiles  --
            Regular Field  Sample  (QC Code  field is blank) 	 47
      1.12  Record Type 20s and 30s for Volatiles  --
            Field Sample chosen for MS/MSD (QC Code  'LSD')   ...  	 48
      1.13  Record Type  20s and 30s  for Volatiles  --
            Matrix Spike Sample  (QC Code  'LF1')  	 48
      1.14  Record Type  20s and 30s  for Volatiles  --
            Matrix Spike Duplicate  Sample  (QC  Code 'LF2') 	 50

2.0   SEMIVOLATILES	52
      2.1   Record Type 10 for Semivolatiles	52
      2.2   Record Type 11 for Semivolatiles	52
      2.3   Record Type 20s and 30s  for Semivolatiles --
            Initial Calibration Mean Values  (QC Code  'MNC1)	52
      2.4   Record Type 20s and 30s for Semivolatiles --
            DFTPP Tune (QC Code  'LPC')	52
      2.5   Record Type 20s  and 30s for Semivolatiles --
            Initial Calibration Standard  (QC Code  'CLM')  	 52
      2.6   Record Type 20s  and 30s for Semivolatiles --
            Dual Purpose Calibration Standard  (QC  Code 'CLD')  	 53
      2.7   Record Type 20s  and 30s for Semivolatiles --
            Continuing Calibration  Standard  (QC Code  'CLC1)  .  .  ...  .  .  .  .54
      2.8   Record Type 20s  and 30s for Semivolatiles --
            Method Blank (QC  Code  'LRB')	55
      2.9   Record Type 20s  and 30s for Semivolatiles --
            Regular  Field  Sample  (QC Code field  is blank)  	 56
      2.10  Record Type 20s  and 30s for Semivolatiles --
            Field Sample Chosen for MS/MSD (QC Code 'LSD')   	57
                                      H-39                              OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
      2.11  Record Type 20s and 30s for Semivolatiles --
            Matrix Spike Sample  (QC Code  'LF1')	 58
      2.12  Record Type 20s and 30s for Semivolatiles --
            Matrix Spike Duplicate Sample  (QC Code  'LF2')  	 59

3.0   PESTICIDES  (COL. 1)	62
      3.1   Record Type 10 for Pesticides  (Col 1)	62
      3.2   Record Type 11 for Pesticides  (Col 1)	62
      3.3   Record Type 20s and 30s for Pesticides  (Col 1)  --
            Initial Calibration Mean Values  (QC  Code  'MNC')  ........ 62
      3.4   Record Type 20s and 30s for Pesticides  (Col 1)  --
            Florisil Cartridge Check Recovery Values  (QC Code  'FLO')   ... 63
      3.5   Record Type 20s and 30s for Pesticides  (Col 1)  --
            GPC Recovery Values  (QC Code  'GPC1.)	.  . 64
      3.6   Record Type 20s and 30s for Pesticides  (Col 1)  --
            Resolution Check Standard  (QC Code 'LPC') 	 64
      3.7   Record Type 20s and 30s for Pesticides  (Col 1)  --
            Performance Evaluation Mixture  (QC Code  'CLE')   	 65
      3.8   Record Type 20s and 30s for Pesticides  (Col 1)  --
           . Initial Calibration Multicomponent Standard  (QC Code  'CLS')  .  . 66
      3.9   Record Type 20s and 30s for Pesticides  (Col 1)  --
            Initial Calibration Single Component --
            Individual Standard A  (QC Code  'CLM')	67
      3.10  Record Type 20s and 30s for Pesticides  (Col 1)  --
            Initial Calibration Single Component --
            Individual Standard Mix B  (QC Code 'CLM')	68
      3.11  Record Type 20s and 30s for Pesticides  (Col 1)  --
            Instrument Blank  (QC Code  'LIB')   	69
      3.12  Record Type 20s and 30s for Pesticides  (Col 1)  --
            Method Blank  (QC Code  'LRB')   	70
      3.13  Record Type 20s and 30s for Pesticides  (Col 1)  --
            Matrix Spike Sample  (QC Code  'LF1')	70
      3.14  Record Type 20s and 30s for Pesticides  (Col 1)  --
            Matrix Spike Duplicate Sample  (QC Code  'LF2')  	 71
      3.15  Record Type 20s and 30s for Pesticides  (Col 1)  --'
            Continuing Performance Check  --
            Individual Standard Mix A  (QC Code 'CLE')	73
      3.16  Record Type 20s and 30s for Pesticides  (Col 1)  --
            Continuing Performance Check  --
            Individual Standard Mix B  (QC Code 'CLE')	73
      3.17  Record Type 20s and 30s for Pesticides  (Col 1)  --
            Continuing Performance Check  --
            Performance Evaluation Mixture  (QC Code  'CLE')  c	 74
      3.18  Record Type 20s and 30s for Pesticides  (Col 1)  --
            Field Sample chosen for MS/MSD  (QC Code  'LSD')	75

4.0   PESTICIDES  (COL. 2)  .	77
      4.1   Record Type 10 for Pesticides  (Col 2)	77
      4.2   Record Type 11 for Pesticides  (Col 2)	77
      4.3   Record Type 20s and 30s for Pesticides  (Col 2)  --
            Initial Calibration Mean Values  (QC  Code  'MNC')	77
      4.4   Record Type 20s and 30s for Pesticides  (Col 2)  --
            Resolution Check Standard  (QC Code 'LPC') 	 78
                                     H-40                              OLM04.2

-------
                                                       Exhibit H -- Appendix A
                                           Format of Records for Specific Uses
      4.5    Record Type 20s and 30s for,Pesticides (Col 2)  --
            Performance Evaluation Mixture (QC Code 'CLE')   	 79
      4.6    Record Type 20s and 30s for Pesticides (Col 2)  --
            Initial Calibration Multicomponent Standard (QC Code 'CLS') .   . 80
      4.7    Record Type 20s and 30s for Pesticides (Col 2)  --
            Initial Calibration Single Component --
            Individual Standard A (QC Code 'CLM')	.80
      4.8    Record Type 20s and 30s for Pesticides (Col 2)  --
            Instrument Blank (QC Code 'LIB')   	 82
      4.9    Record Type 20s and 30s for Pesticides (Col 2)  --
            Method Blank (QC Code 'LRB')   	82
      4.10  Record Type 20s and 30s for Pesticides --
            Matrix Spike Sample (QC Code 'LF1')	83
      4.11  Record Type 20s and 30s for Pesticides --
            Matrix Spike Duplicate Sample (QC Code 'LF2')  	 84
      4.12  Record Type 20s and 30s for Pesticides (Col 2)  --
            72 Hours Aroclor Standard (QC .Code 'CLE')	86
      4.13  Record Type 20s and 30s for Pesticides (Col 2)  --
            Field Sample chosen for MS/MSD (QC Code  'LSD')    	86

5.0   TYPE 90 -- COMMENTS RECORD	88
                                     H-41                               OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
1.0   VOLATILES

1.1   Record Type 10 for Volatiles

      10||||||GC/MS|OLM04.2V\ \TESLAB||||68D00001|INSTV1||TEST LABS
      INC.||000012C33

1.2   Record Type 11 for Volatiles

      11|DB-5| |0.53| | |  | | | | | | | |000024CD9

1.3   Record Type 20s and 30s for Volatiles -- Initial Calibration Mean Values
      (QC Code  'MNC')

      20|||0|MNC||18000|X1201|||||||||36|||000036668

      30|C|74873|||AVG|0.53l||||||R|17.0||||||||||||||0000486BE

      30|C|74839|||AVG|l.536||||||R|11.1||||||||||||||00005A717
1.4   Record Type 20s and 30s for Volatiles -- BFB Tune  (QC Code  'LPC')

      20|||0|LPC||18000|X1201|1532|04|25|01|01||||l|||0004011E3

      21|N||||||BFB01|||||| MM |0004121B2

      30|C|460004||||||||||||||||||||MII|00042388C

      36|M| | |50| 118.8|75|45.5| |95|100.0| |96|7.2| 1173|0.0|0.01174|72.8| 1175 J5.7
      |7.8|176|71.6|98.3|177|4.9|6.8|MM  M M M M|000439122

1.5   Record Type 20s and 30s for Volatiles -- Initial Calibration Standard  (QC
      Code  'CLM')

      20 | | VSTD020|0|CLM| 118000|X120111992103|25|09|42| M I36M |00044B8F2

      2i|N|| M  IISTDOIJ M M M M I|ooo45C944

      23 I P I 1992 |.04 | 25 | 011 011 BFB01 M M M M M M  M M M M M M M 00046EA64

      30 I C| 74873 I 74975 I | |0.563| M M M M M M M  M M I | 000470F21



                                      H-42                              OLM04.2

-------
1.6
 1.7
      30 IC|74839 I 74975 I | 11.528|
                                                       Exhibit H -- Appendix A
                                           Format of Records for Specific Uses
                                              I0004833E2
 Record Type 20s and 30s for Volatiles -- Dual Purpose Calibration Standard
  (QC  Code  'CLD')

 20||VSTD050|0|CLD||18000|X1201|1992\05|25|10|25||||39|| J0014300AA

. 21|N||||||CSTD01|||||||||||00144117D

 23|P|15S2|05|30|05|01|BFB02|| | | | | | | I 1 | | | | I ! M I I I ! 1 I I I0014532A2

 30|C|74873|74975|||0.931|||||||||||||||||||||001465B66

 30|C|74839|74975| | |l.82l| | | | | I I I I I I I I I I I I I I I I00147842C
  30|C|17060070|74975| | |2.089|
                                                      | 001810B47
  30|C|3114554| | | STD | | |A|50.0| || I I I I I I I I I I I I 1 1 1 109281 j 001822EDB

  32| | |RT|15.74| | | | | | | | | | | 001833EF5

  30 | C| 5403 63 | | | STD | | | A| 50 . 0 | | | | | | | | . | | | |

  32| | |RT|9.97| | | | | | | | | | (0018570D13

  30 | C| 74975 | | | STD | | | A| 50 . 0 | | | | | | | | | | | |

  32 | | |RT|8.00| | | | | | | | | | |00187A10C
                                                 1 125103  00184618CB
                                               | I | 26084 | 0018691EDC
  Record Type 20s and 30s for Volatiles -- Continuing Calibration Standard
  (QC Code 'CLC1 )

  20| |VSTD050|0|CLC| 1 18000 |X1201 1 1992 \ 05 | 30 | 05 | 25 | |.| |39| |.|0024300AD

  2l|N| | | | | JCSTDOI) | | | | |  | | | | |00244117F
                                      H-43
                                                                        OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
      23|P|1S52|05|30|05|01|BFB02| | || I I I


      30 |C| 74873 | 74975 | | |0.96l| | |  | | |  | | | |


      30 | C| 74839 | 74975 | | |l.803| | |  | | |  | | | |
1.8
                                               I I M I I I I! I I I |0024532A1


                                               | F| 81 . 0 | | | 002465B65
                                               | F| 17 . 4 | | |00247842A
                                        I I I I I I I  I I |F|l0.6| | |002810B46


                                        | I I I I I I  I 1 1 1 109281 1 002822EDA
                                                | I | 125103 | 00284618C
      30|C|17060070|74975| | |2.089| | | | || I


      30|C|3114554| | |STD| | |A|50.0| | | | | | |


      32 | | |RT| 15. 74) | | | | | | | | | |002833EF3


      30 | C| 540363 | | |STD| | | A| 50 . 0 | | | | | | | |
                                                           '

      32| | |RT|9.97| | | | | | | | | | |0028570D1


      30|C|74975| | | STD | | |A|50.0| | | | | | | | | || I I I I 1 1 1 26084 | 0028691ED


      32| | |RT|8.00| | M I I I I I I |00287A10D

      Record Type  20s and 30s  for Volatiles  --  Method Blank  (QC Code  'LRB')

      20| |VBLK01|1|LRB| | 18000 |X1201 | 1992\ 05 | 30 | 06 | 11 | |ML|5.0|39| | | 00288CB9F


      2l|N|L| | | | |FBLK01| | | I I I I I I I | 00289DCF6


      22|l$52|05|30|05|25|CSTDOl| | | | |l.0| | | |00290F6D3


      23|P|l552|05|30|05|Ol|BFB02| | | | | | | || | | I I I I I I I p I ° I T I ° I I I I | 002911B17


      30 | C| 74873 | 74575 | UG/L | BDL | |u| | |u| 10 | | | | | | | | | | | | | | | | 002923881


      30|C|74839| 74575 1 UG/L | BDL | |u| | |U|lO| | | | | | I I I I I I I I I I 0029355ED
30 | C| 2037265 | 3114554 | UG/L| | 47 . 9 | | A| 50 . 0 |


30|C|460004|3114554|UG/L| | 47 . 0 | | A| 50 . 0 | |
                                                     | | S | 96 |


                                                      S 94
                                                                 | | 003251E4B


                                                                  0032647E1
                                     H-44
                                                                       OLM04.2

-------
                                                       Exhibit H --  Appendix A
                                           Format of Records for Specific Uses
      30|C|17060070|74975|UG/L| | 50 . 3 | | A| 50 . 0 | | | | | | | |.| |S|lOl| | | | | | | 003277241


      30|C|3114554| | | STD | | |A|50.0| | | | | | | | | | | | | | | | I | 104439 | 0032895DF


      32 | | | RT | 15. 70 | | | | | | | | | | |00329A5EA

      30 | C| 540363 | | | STD | | | A| 50 . 0 |  | | | | | | | | | | | | | | | I | 135938 | 00330C8A8


      32| | |RT|9.92| | | | | | | | | | | 00331D7DE


      30|C|74975| | | STD | | |A|50.0| |  | | | | | | || II I I I | 1 1 26488 | 00332F90C


      32| | |RT|8.00| | | | | | | | | | |00333082C

1.9   Record Type 20s and 30s for Volatiles -- Storage Blank (QC Code 'LHB')

      20 | |VHBLKOl|l|LHB| | 18000 |X1201 | 1592 | 05 | 30 | 06 | 21 | |ML|5.0|39| | | 00688CB9F


      2i|N|L| 1 1 1 IFHBLKOI) || M  1 1  I I  I |oo689DCF6

      22 | 1992 | 05 | 30 | 05 | 25 | CSTD01 1  | | | | 1 . 0 |  | |  | 00690F6D3


      23 | P| 1952 | 05 | 30 | 05 | 01 | BFB02 | | | | | | | |  | |  |MB | 1992 | 05 | 30 | 06 | 11 | FBLK01 | P | 0 | T | 0
      |  | | | |006911B17


      30 | C| 74873 | 74975 |UG/L |BDL | |U| | |u| 10 | |  | | | | | | | | | | | |  | | 006923881


      30|C|74839| 74975|UG/L | BDL | |u| | |u|lO| || I I I II I I I I I I  I I 0069355ED
      30|C|2037265|3114554|UG/L| |47.9| |A|50.0| | || | I I I I IS|96| | | | I I I007251E4B


      30|C|460004|3114554|UG/L| | 47 . 0 | |A|50.0| | | | | | | | | |S|94| | | | | | |0072647E1


      30|C|17060070|74975|UG/L| |50 .3 | |A|50.0| | | | | | | | | |S1101| | | | | | |007277241


      30|C|3114554| | |STD| | |A|50.0| | | | | I I I I I I I I I I 111104439|0072895DF


      32|||RT|15.70|||||||||||00729A5EA


      30|C|540363|||STD|||A|50.0||||||||||||||||I|135938|00730C8A8


      32|||RT|9.92|||||||||||00731D7DE

                                      H-45                               OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses


      30|C|74975| | |STD| |  |A|50.0| || I Ij I I I M I I M 11|26488|00732F90C

      32|||RT|8.00|||||||||||00733082C

1.10  Record Type 20s and 30s for Volatiles -- Instrument  Blank (QC Code  'LIB')

      20||VIBLKOl|l|LIB||18000|X1201|1992\05|30|06|30||ML|5.0|39|||00688CB9F

      21|N|L| | |  | |FIBLK01| | | I I I I I I I |00689DCF6

      22|1992|05|30|05|25|CSTD01| | | | |l.0| | | |00690F6D3

      23|P|1992|05|30|05|01|BFB02| | | | || I I I I I MB\1992\05|30|061111FBLK01|P|0|T|0
      I  I I I I006911B17
      30|C|74873|74975|UG/L|BDL||u|||u|lO|||||||||||||||

      30|C|74839|74975|UG/L|BDL||u|||u|10||||||||||||||||0069355ED
      30|C|2037265|3114554|UG/L||47.9||A|50.0||||||||||S|96|||||||007251E4B

      30|C|460004|3114554|UG/L||47.0||A|50.0||||||||||S|94|||||||0072647E1

      30|C|17060070|74975|UG/L| |50.3| |A|50.0| || | I I I I I ls|!0l| I I I I I |007277241

      30|C|3114554| | |STD| | |A|50.0| | || I I I I I I I I I I I 111104439|0072895DF

      32 | |  |RT|15.70| |  | | | | | | |,| |00729A5EA

      30|C|540363| | |STD| | |A|50.0| | | I I I I I I I I I I I I 111135938|00730C8A8

      32|||RT|9.92|||||||||||00731D7DE

      30|C|74975| | |STD| | |A|50.0| | | | | | || I I I I I I I 11126488|00732F90C

      32| |  |RT.|8.00| | |  | | | | | | | |00733082C
                                     H-:46                              OLM04.2

-------
                                                       Exhibit H -- Appendix A
                                           Format of Records for Specific Uses
1.11  Record Type 20s  and 30s for Volatiles -- Regular  Field Sample  (QC Code
      field is blank)

      20| |X1200|l| | |18000|X1201|1992|05|30|06|37| |ML|5.0|39| | | 0033432DB


      21 |N | L | | | | |FX1200| | | | 1 1592 | 05 | 27 I | | |003354C2D


      22 | 1992 | 05 | 30 | 05 | 25 | CSTD01 | | | | |l.0| | | |00336660A


      23|P|1992|05|30|05|01|BFB02|HB|1992|05|30|06|21|FHBLK01| | | |MB | 1992 | 05 | 30
      |06|ll|FBLKOl|P|0|T|2| | | | 0 | 003379931
30 | C| 74873 | 74975 |UG/L |BDL | |U| | |U| 10 |


30|C|74839| 74975 |UG/L| BDL | |U| | |U| 10 |
                                                         J00338B69B


                                                         00339D407
30|C|2037265|3114554|UG/L| | 51 . 8 | | A| 50 . 0


30|C|460004|3114554|UG/L| |59.0| |A|50.0|


30|C|17060070|74975|UG/L| | 54 . 3 | |A|50.0|


30|C|3114554| | | STD |  | |A|50.0| | |  | | |  I I I  I I  I

32| | |RT|15.66| | | |  | |  | | | | [00275277B


30 |C| 540363 |  | | STD | |  | A| 50 . 0 |  | | |  | | |  | | |  | |  |


32| | |RT|9.91| | | | |  | |  | | | [002775948


30 |C | 74975] |  | STD |  | | A| 50 . 0 | |  |'| |  | | |  | | |  | |  |


32| | |RT|7.96| | | | |  | |  | | | J00279898B


30|C| |  |UG/L|TIE|9| |  | | | | | | | |  | | |  | | |  | ||  I |002807A4A
                                                | | S | 102 |


                                                |S|98| | |

                                                |S|l07| |
                                                                 | | | 00271A07D


                                                                 |00272CA19

                                                                 | | 00273F48C
                                                  I  96179  002741762
                                                | I | 123502 | 002764A14
                                               | 1 1 26331 1 002787A4A
32| | |RT|2.90|
                             |00281898B
 33 | UNKNOm\ 00281898B


 30|C|74630527| |UG/L | TIE 1 1 1 JN|
                                                      | 002827A4.A
                                      H-47
                                                                        OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses


      32| | |RT|25.20| | | |.| | |  | | | |00283898B

      33|3-UNDECENE, 6-METHYL-,  (E)-|00284898B

1.12  Record Type 20s and  30s for  Volatiles --  Field Sample chosen for MS/MSD
      (QC Code 'LSD1)

      20| |X120l|l|LSD| 118000|X120111552|05|30|06|37 | |ML|5.0|39| | |0033432DB

      21|N|L|||||FX1201LSD||||\1992\05\27\\\|003354C2D

      22|1552|05|30|05|25|CSTD01|||||l.0||||00336660A

      23|P|1552|05|30|05|01|BFB02|HB|1552|05|30|06|21|FHBLK01||||MB|1552|05|30
      |06|11|FBLK01|P|0|T|0||||0|003379931

      30|C|74873|74575|UG/L|BDL||u|||u|10||||||||||||||||00338B69B

      30|C|74839|74575|UG/L|BDL||u|||u|10||||||||||||||||00339D407
      30|C|2037265|3114554|UG/L||50.8||A|50.0||||||||||s|l02|||||||00371A07D

      30|C|460004|3114554|UG/L| |49.0| |A|50.0| | || I I I I I |S|98| | | | | | |00372CA19

      30|C|17060070|74975|UG/L| |53.3| |A|50.0| | | | | | I I I |S|l07| | | | | | |00373F48C

      30|C|3114554.| |  | STD | | |A|50.0| | | | | | I I I I I I I I II * I 96178 | 003741762

      32|||RT|15.65|||||||||||00375277B

      30|C|540363 | | |STD| | |A|50.0| | | | | | | | | | | | | | | 1111235011003764A14

      32|||RT|9.90|||||||||||003775948

      30|C|74975|||STD|||A|50.0||||||||||||||||I|26330|003787A4A

      32|||RT|7.95|||||||||||00379898B

1.13  Record Type 20s  and 30s  for Volatiles --  Matrix  Spike Sample  (QC Code
      1LF1')

      20||X1201MS|1|LF1||18000|X1201|1992\05|30|07|14||ML|5.0|39|||00380B5C1


                                     H-48                              OLM04.2

-------
                                                 Exhibit H -- Appendix A
                                     Format of Records for Specific Uses
2l|N|L| | | | |FX1201LF1| | | | \1992\QS\21\ | | |00381CF38


22 | 1992 | 05 | 30 | 05 | 25 | CSTD01 | | | | |l.0| | | |00382E915


23 | P | 1992 | 05 | 30 | 05 | 01 1 BFB02 | HB | 1992 1 05 | 30 | 06 | 21 j'FHBLKOl | | | |MB 1 1992 1 05 | 30
|06|11|FBLK01|P|0| | |S|0|R|0|003831CEC
30|C|74873| 74 795 \ UG/L | BDL |
                                                   003843A56
30|C|75354|74975|UG/L||40||A|50.00|||||P|80||D|4|||||||||003911212




30|C|79016|540363|UG/L||46||A|50.00|||||P|92||D|4|||||||||004026477




30|C|71432|540363|UG/L||46||A|50.00|||||P|92||D|2|||||||||00405C84B




30|C|108883|3114554|UG/L||45||A|50.00|||||P|9l||D|6|||||||||00412ABC4


30|C|108907|3114554|UG/L||44||AJ50.00|||||P|88||D|7|||||||||00413D5FF
30|C|2037265|3114554|UG/L| |46.l| |A|50.0|


30|C|460004|3114554|UG/L| | 46 . 7 | | A| 50 . 0 | |


30 | C| 17060070 | 74975 | UG/L | | 54 .3 | | A| 50 . 0 | |
                                                | | S | 92 | | | | | | |004175BA7


                                                | S | 93 | | | | | | | 004188542


                                                |S|l09| | | | | | J00419AFC2
30|C|3114554| | |STD| | |A|50.0|
 32   |RT|15.70
                         00421E35B
 30 |C| 540363 | | | STD | | | A| 50 . 0 | | | | | | |


 32| | |RT|9.92| | | | | | | | | | |004231549


 30|C|74975| | | STD | | |A|50.0| | | | | | | |


 32| | |RT|7.98| | | | | | | | | | |00425459B
                                     I I I I I I 1 1 1 101076 | 00420D350
                                          | I | 125573 | 004220613
                                            26701 004243657
                                H-49
                                                                  OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
1.14  Record Type 20s and 30s for Volatiles -- Matrix Spike Duplicate Sample  (QC
      Code ' LF2 ' )

      20| |X1201MSD|l|LF2| | 18000 |X1201 | 1992 \ 05 | 30 | 07 | 55 | |MLJ5.0|39| | | 00426726B


      2l|N|L| | | | |FX1201LF2 |  | | | \1992\ 05 |27| | |  J004278BE3
22 | 1992 | 05 | 30 | 05 | 25 | CSTD01
                                     l.0    00428A5CO
      23 | P | 1992 1 05 | 3 0 | 05 | 01 | BFB02 | HB 1 1952 1 05 | 30 | 06 | 21 1 FHBLK01 | | | | MB | 1992 1 05 | 3 0
      |06|ll|FBLKOl|P|0| | |S|0|R|0|00429D997
30 | C| 74873 | 74975 |UG/L |BDL|
                                 10 |
                                                         00430F701
      30|C|75354|74975|UG/L||38||A|50.00|||||PJ|77|D|4||||




      30|C|79016|540363|UG/L||44||A|50.00|||||P||88|D|4||j




      30|C|71432|540363|UG/L||47||A|50.00|||||P||94|D|2|||




      30|C|108883|3114554|UG/L||48||A|50.00|||||P||97|D|6|


      30|C|108907|3114554|UG/L||47||A|50.00|||||P||94JD|7|
                                                        [00437CED4




                                                        ||004482146




                                                        ||004518527




                                                        ||| J0045868B3


                                                        I I I I0045992EE
                                               | |S|98| | | | | | |0046318AA


                                               | S | 95 |  | | | | | | 004644251


                                                            004656CBF
      30|C|2037265|3114554|UG/L| |49.2| |A|50.0| | | | |


      30 | C| 460004 | 3114554 | UG/L |  | 47 . 6 | | A| 50 . 0 | | | | | |


      30|C|17060070|74975|UG/L|  | 55 . 2 | |A|50.0| | | | | |


      30|C|3114554| | | STD | | |A|50.0| | || I I I I  I I I  I I I I | I 1 101145 | 00466904A


      32| |  |RT|15.70| | |  | | | | |  | | |00467A055


      30 |C| 540363 | | | STD | | | A| 50 . 0 |  | | |  | | | | |  | | |  | | | 1 1 1 124184 | 00468C30A

                                     H-50
                                                                       OLM04.2

-------
                                                 Exhibit H -- Appendix A
                                     Format of Records for Specific Uses
32|||RT|9.94||||||||||(00469D24C


30|C|74975| | |STD| | |A|50.0| | | 1 | 1 | | | I II I I I |1126683|00470F36D


32RT7.980047102B1
                                H-51                              OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
2.0   SEMIVOLATILES

2.1   Record Type 10 for Semivolatiles

      10||||||GC/MS|OLM04.2B\\TESLAB||||68D00001|INSTB1||TEST LABS
      INC.||000002BA9

2.2   Record Type 11 for Semivolatiles

      ,ll|XRT-l||0.52||||||||||J000024C8D

2.3  ' Record Type  20s  and  30s  for Semivolatiles  --  Initial  Calibration Mean
      Values (QC Code  'MNC')

      20|||0|MNC||18000|X1201|||||||||72|||00003661C

      30|C|108952|3855821||AVG|l.817||||||R|7.2||||||||||||||000048677

      30|C|111444|3855821|  |AVG|l.607|  | | | |  | R | 6 .2 | | |  | | | | | | | | | | | 00005A6C4

      30|C|95578|3855821||AVG|l.375||||||R|3.l||||||||||||||00006C650
      30|C|93951736|3855821||AVG|l.22l|||MIRI5-3IIIIIIIIIIIIII000745BDE

      30|C|2199691|3855821||AVG|0.91l||||||R|2.0||||||||||||||000757DOA

2.4   Record Type 20s and 30s for Semivolatiles -- DFTPP Tune  (QC Code 'LPC')

      20|||0|LPC||18000|X1201|1552|05|01|00|32||||l|||000769E6E

      21|||||||DFTPP01|||||||||||00077AF43

      30|C|5074715|||||||||||||||||||||||||0007BC71C

      36|M| | | 511 |41.9|68 | 0.5 11.3|69|37.6| |70|0.0|0.01127|45.2| |l97|0.4| |l98|lO
      0.0| 1199|8.0| |275|25.0| |365|3.69| |44l|l3.7| |442J99.l| |443| | 20.8|21.0|000
      793554

2.5   Record Type 20s and 30s for Semivolatiles -- Initial Calibration Standard
      (QC  Code  'CLM')

      20||SSTD080|0|CLM||18000|X1201|1992\05|01|02|01||||72|||00155B94D

      21|||||||STD02|||||||||||00156C8F3


                                      H-52                              OLM04.2

-------
                                                       Exhibit H --  Appendix A
                                           Format of  Records for Specific  Uses
      23 | P |1992|05|01100|32|DFTPP011 | | || I I I I I I I I I I I I I I I I I II I I00157EBBA


      30|C|108952|3855821| | |l.853| | | | || I I I I I I I I I I I I I I I0015812F5  .-


      30|C|11144438558211.599' | |  |                 001593A37
      30|C|93951736|385582l| | |l.248J |A|40.000| | || I II II I I I I I I I I |00228C51D


      30|C|2199691|3855821| | |0.928| |A|40.000| | | | | I I I I I I I I I II I (00229ED4C

2.6   Record Type  20s  and  30s  for Semivolatiles  --  Dual Purpose  Calibration
      Standard (QC Code 'CLD')

      20||SSTD050|0|CLD|| 18000|X120111992|05|01102|40||||72|||0045914F6


      211 | | | | | |STD03| |  | | | |  | | | | |00460249D


      23|P|1992|05|01|00|32|DFTPP01| || I I I  I I  I I I I I I I I I I I I I I I I I I 004614764


      30|C|108952|3855821|  | |l.462| | |l  I I I I  I I  I II I I I I I I I |004623B3A


      30|C|111444|3855821|  | |l.233| || I  I I M  II  I I I I I I M I I |00463664C
      30|C|2199691|3855821| | |0.828| | | || I I I I I I I I I I I I M I I0053318FO


      30|C|1146652| | |STD| | |A|20.0| | || I I I I I I I I I I I |I|58474|005343BC2


      32|||RT|7.23|||||||||||005354AE6


      30|C|15067262| | |STD| | |A|20.0| || I I I I I I I I I I I I |I|27242|005366E6F


      32|||RT|9.17||||||||||(005377DA2


      30|C|1517222| | |STD| | |AJ20.0| | || I I I I I I I I I I I 11|38472|00538A043


      32| | |RT|10.82| | | M | I I M |00539B042


      30|C|1520963| | | STD | | |Al20•0| | I I I I I I I I I I I I I 11118770|00540D2F2

                                      H-53                              OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses


      32|||RT|17.95|||||||||||00541E310

      30|C|171903.5| | |STD| |  |A|20.0| | | | |  | | | | | | I I I I II I 24292 | 0054205B2

      32|||RT|14.22|||||||||||0054315AF

      30|C|385582l| | |STD| |  |A|20.0| | | | I  I I I I I I I I I I 11 117522|00544385F

      32|||RT|5.87|||||||||||00545479F

2.7   Record  Type  20s  and 30s  for Semivolatiles --  Continuing  Calibration
      Standard  (QC  Code  'CLC')

      20||SSTD050|0|CLC||18000|X1201|1992\06|01|11|30||||78|||00459DC7F

      21|||||||CCSTD01|||||||||||00460ED24

      23|P|lS32|06|Ol|ll|Ol|DFTPP02| | |  | | | I | | | | | | | |  I I  | I I | | |  | | [004610FEB

      30|C|108952|3855821|  | |l.462| | | | |  | | | | | | I I I | I |F|l9.5| |  [004623B3A

      30C1114443855821|l.233F|23.300463664C
      30|C|2199691|3855821| | |0.828| | || I I I I I I I I I I I I IFI9-1| I |0053318FO

      30|C|1146652| | |STD| | |A|20.0| | | | | | || I I I I I I I |I|58474|005343BC2

      32|||RT|7.23|||||||||||005354AE6

      30|C|15067262| | |STD| | |A|20.0| | || I I I I I I I I I I I |1127242|005366E6F

      32|||RT|9.17|||||||||||005377DA2

      30|C|1517222| | |STD| | |A|20.0| | | | | | | | | | | | | | | |I | 38472|00538A043

      32|||RT|10.82|||||||||||00539B042

      30|C|1520963 | | |STD| | |A|20.0| | | | | | | | | | | | | | | 11118770|00540D2F2

      32|||RT|17.95|||||||||||00541E310

      30|C|1719035| | |STD| | |A|20.0| | | | | | | | | | | | | I I 11|24292|0054205B2

                                     H-54                              OLM04.2

-------
                                                       Exhibit H --  Appendix A
                                           Format of Records for Specific Uses
32 | | |RT|14.22| | | | | | | | | | |0054315AF


30|C|385582l| | | STD | | |A|20.0| | | | | | |
                                                  1 17522 | 00544385F
      32| | |RT|5.87| | | | | | | | | | J00545479F

2.8   Record Type 20s  and  30s  for Semivolatiles  -- Method Blank  (QC Code 'LRB')

      20| |SBLK32|l|LRB| | 18000 |X1201 | 1552| 06 | 01 | 12 | 10 | |ML|lOOO|78| | |005467308


      21 1 |L| |C| |  |FSBLK01|1952|03|30| | | | |  |2.0| | 005478CBB


      22 | 1552 | 06 | 01 | 11 | 30 | CCSTD01 | | | |lOOO|l.O| |  | | 00548AA6D
23|P|l952|06|Ol|ll|Ol|DFTPP02| | | | | | | | | |


30 | C| 108952 | 3855821 1 UG/L | BDL| |u| | |u| 10 |
30 | C| 4165600 | 1520963 | UG/L | | 40 . 9 | | A| 50 . 0 | | |


30|C|321608|15067262|UG/L| |42.8| |A|50.0| | |


30|C|98904439|1719035|UG/L| | 52 . 0 | |A|50.0| |


30|C|4165622|3855821|UG/L| |67.2| |A|75.0| | |


30 |C| 367124 | 3855821 1 UG/L | | 61 . 7 | | A| 75 . 0 | | |


30|C|118796|15067262|UG/L| | 70 . 9 | |A|75.0| |


30|C|93951736|3855821|UG/L| |69.2| |A|75.0|


30|C|219969l|385582l|UG/L| |38.9| |A|50.0| |


30 | C| 1146652 |  | | STD | | | A| 20 . 0 |  | | | | | |  | | | | | | |
 32| | |RT|7.72|
                             |00623CF6A
       30 C  15067262    STD   A 20 . 0
                                                     |P|0|T|0| | |  | | 00549D056


                                                        | | | | | 00550EE8E
                                                 S | 82 |
                                                 S  86
                                                  |S|l04|
                                                  S  90
                                                 S  82
                                                  S  95
                                                  | S | 92
                                                  S  78
                                                                  006147181
                                                                  | 006159BF9
                                                                    | 00616C83A
                                                                  | 00617F2D2
                                                                 |006181C81
                                                                  | 00619472C
                                                                   0062072BF
                                                                  | 006219D66
                                                  I  94564  00622C038
                                                   I  46152  00624F2FE
 32| | |RT|9.65|
                             | 00625023E
                                H-55
                                                                        OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
      30 | C| 1517222 | | | STD | | | A| 20 . 0 |
2.9
      32 |  | RT | 11 . 32
                              0062734DF
      30|C|1520963| | | STD | | |A|20.0| | | | | |  |


      32| | |RT|18.44| | | | | || I  I I I 0062967B8


      30|C|1719035| | |STD| | |A|20.0| | | | | |  |


      32 | | | RT | 14. 70 | | | | | | | |  | | |006319A71


      30|C|3855821| | | STD | | |A|20.0| | | | | |  |


      32.| | |RT|6.35| | || | | | | |  | | 00633CC63
                                                 | I | 65936 | 0062624EE
                                                   | 35768 | 00628579F
                                                   43708 006308A67
                                                   1 29752 | 00632BD33
      Record Type 20s and 30s for Semivolatiles -- Regular Field Sample (QC Code
      field is blank)

      20| |X120l|l| | |18000|X1201|15S2|06|01|12|20| | ML 1 1000 | 78 | | | 00434F7E4


      21| |L| |C| | |FX1201|1SS>2|05|30| \1992\05\27\ \2.0\ | 0043518E4


      22 | 1992 | 06 | 01 | 11 | 30 |CCSTD01 | | | |lOOO|l.O| | | |004363696


      23 | P | 1992 1 06 | 01 1 11 | 01 1 DFTPP02 | | | I I I I I I I I MB I 1992 | 04 | 01 1 15 1 14 | FSBLK01 1 P | 0 |
      T|0| | | |0|004376C26
      30 | C| 108952 | 3855821 |UG/L|BDL| |U| | |U|10|
                                                             004388A5E
30|C|4165600|1520963|UG/L| | 43 . 0 | | A| 50 . 0 | |


30 | C| 321608 | 15067262 | UG/L| | 48 . 3 | | A| 50 . 0 | |


30|C|98904439|1719035|UG/L| |51.8| |A|50.0|


30 | C| 4165622 | 3855821 | UG/L| | 69 . 8 | | A| 75 . 0 | |


30 |C| 367124)3855821 | UG/L |  | 69 . 6 | | A| 75 . 0 | | |


30|C|118796|15067262|UG/L| |79.5| |A|75.0| |



                               H-56
                                                       | S 86
                                                       | S | 97 |
                                                   I I I I I |S|l04|
                                                       S 93
                                                      | S | 93 |
                                                       |S|l06|
                                                                   005020B12
                                                                   00503358D
                                                                    0050461DF
                                                                  005058C8C
                                                                 | 00506B64E
                                                                   | 00507E1D3
                                                                       OLM04.2

-------
                                                       Exhibit  H -- Appendix A
                                           Format  of  Records  for Specific  Uses
30 |C| 93951736 | 3855821 |UG/L| | 73 . 0 | | A| 75 . 0 |


30|C|219969l|385582l|UG/L| |45.6| |A|50.0| |


30|C|385582l| | | STD | | |A|20.0| |  | |  | | | |  | |  | |  | |


32| |  |RT|6.38| | | |  | | | | | |  |0051169E8


30|C|1146652| | | STD | | |A|20.0| |  | || | | |  | |  | ||


32| |  |RT|7.75| | | |  | | | | | |  |005139BCF


30|C|15067262| | |  STD | | |A|20.0|  | |  | | | |  | |  | |  |
                                                 | S | 97 |
                                                 S 91
                                                                  | 005080D64
                                                                 |005093801
                                                   30288 | 005105AB5
                                                   96100 005128C90
                                                   1 47432 | 00514BF65
32   RT 9 . 67
                             00515CEA7
      30|C|1517222| | | STD | | |A|20.0| | | | | || I I I I I I I I 1 1 1 71076 | 00516F145


      32| | |RT|11.34| | M I M|] ||005170142


      30|C|1719035| | |STD| | |A|20.0| | | | | | I I I I I I I II | I | 36728 | 0051823F5


      32| | |RT|14.74| ||  M I I I  I I |00519340D


      30 | C| 1520963 | | |STD| | | A| 20 . 0 | | | | | | | | | | | | | | | | I | 27356 | 0052056BD
32   RT 18. 49
                             | 0052166DB
2.10  Record Type  20s and  30s  for Semivolatiles  --  Field Sample  Chosen for
      MS/MSD (QC Code  'LSD')

      20| |X120l|l|LSD| |18000|X1201|1992|06|01|12|50| |ML|lOOO|78| | | 00634F7E4


      21| |L| |C| | |FX1201LSD|1992|05|30| |l992|05|27| |2.0| |0063518E4


      22 | 1992\ 06 | 01 1 11 1 30 | CCSTD01 1 | | |lOOO|l.O| | | |006363696
23 | P | 1992 | 06 | 01 | 11 | 01 | DFTPP02 |
T|0| | | |0|006376C26
                                              | MB | 1992 | 04 | 01 15 14 FSBLK01 P 0
      30|C|108952|3S55821|UG/L|BDL| |u| | |u|lO| M I I I I I I M I I I I I 006388A5E
                               H-57
                                                                       OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
30|C|4165600|1520963|UG/L| |43.0| |A|50.0| |


30|C|321608|15067262|UG/L| |48.3| |A|50.0| |


30|C|98904439|1719035|UG/L| | 51 . 8 | |A|50.0|


30|C|4165622|3855821|UG/L| |69.8| |A|75.0| |


30 | C| 367124 | 3 855821 | UG/L |  | 69 . 6 |  | A| 75 . 0 |  | |


30|C|118796|15067262|UG/L| |79.5| |A|75.0| |


30|C|93951736|3855821|UG/L| | 73 . 0 | | A| 75 . 0 |


30|C|219969l|385582l|UG/L| |45.6| |A|50.0| |


30|C|3855821| |  | STD | |  |A|20.0| | | |  | | | |  | | | |  | |


32| |  |RT|6.38| |  | |  | || I  I I I 0071169E8


30 | C| 1146652 | |  | STD | |  | A| 20 . 0 | | | |  | | | |  | | | |  | |


32 | |  |RT| 7. 75) |  | |  | | | |  | | |007139BCF


30|C|15067262|  | |STD|  | |A|20.0| | |  | | | |  | | | |  | |
                                                |S|86|
                                                | S | 97 |
                                                        S 104
                                                   93
                                                S 93
                                                | S | 106 |
                                                 | S | 97 |
                                                 S 91
                                                                 J007020B12
                                                                  '00703358D
                                                                    0070461DF
                                                                 | 007058C8C
                                                                 00706B64E
                                                                   | 00707E1D3
                                                                   | 007080D64
                                                                  007093801
                                                   30288 007105AB5
                                                 | I | 96100 | 007128C90
                                                    47432 00714BF65
32   RT | 9 . 67
                             00715CEA7
                                           I 71076 00716F145
                                            | 36728 | 0071823F5
                                           I 27356 0072056BD
      30 | C| 1517222 | | | STD | | | A| 20 . 0 | | | | | | | | | | | | |


      32) |  |RT|11.34| | | | | | | | | | |007170142


      30|C|1719035| | |STD| | |A|20.0| | | | | | | | | | || I


      32 | |  |RT| 14 .74 | | | | | | | | | | | 00719340D


      30 | C| 1520963 | | | STD | | | A| 20 . 0 | | | | | | | | | | | |


      32| |  |RT|18.49| | | | | | | | | | | 0072166DB

2.11  Record Type 20s and 30s for Semivolatiles --  Matrix Spike Sample  (QC Code
      'LF1')

      20 | |X1201MS|l|LFl| 1 18000 |X1201 1 1992 \ 06 | 01 1 13 | 34 | |MLJ500|78| | | 00722931C


      21 | |L| |C| | |FX1201LF1|1352|05|30| \1992\05\2T\ |2.0| | 00723B441

                                      H-58                              OLM04.2

-------
                                                      Exhibit H -- Appendix A
                                          Format of Records for Specific Uses
     22|1352|06|01111130|CCSTD011 | | |50011.0 | | | |00724D125


     23|P|1992|06|01|11|01|DFTPP02|||||||||||MB|1332|06|01|12|10|FSBLK01|P|0|
      ||S|0|R|0|00725074C


     30|C|108952|3855821|UG/L||61||A|75.00|||||P|82||D|2|||||||||007263177


     30|C|111444|3S55S21|UG/L|BDL||U|||U|10||||||||||||||||007274FA5


     30|C|95578|3855821|UG/L||61||A|75.00|||||P|81||D|1|||||||||00728790F
30 | C| 4165600 | 1520963 | UG/L | | 37 . 8 | | A| 50 . 0 |


30 | C| 321608 | 15067262 |UG/L| | 40 . 3 | j A| 50 . 0 |
                                                | S | 76 |


                                                 S 81
                                                                 |0079035AC


                                                                  007916004
      30|C|2199691|3855821|UG/L| |38.4| |A|50.0| | | | | | | | | | S | 77 | | | | | | | 007976162


      30|C|1146652| | | STD | | |A|20.0| | | I I I I I I I I I I I I 1 1 1 101044 | 0079884EO
32| | |RT|7.70|
                            |007999410
      30|C|1719035| | |STD| | |A|20.0| | | | | | | | | | | | | | | | I | 37488 | 008064F42


      32 | | |RT|14.69| | | | |  | | | | | | 008075F5E


      30|C|3855821| | | STD | | |A|20.0| | | | || I I I I I I I II 1 1 1 30512 | 0080881FE


      32| | |RT|6.33| | | | | |  | | | | | 008099122

2.12  Record Type 20s and 30s for Semivolatiles --  Matrix Spike Duplicate Sample
      (QC Code 'LF2')

      20 | |X1201MSD|l|LF2| | 18000 JX1201 | 1992 \ 06 | 01 | 14 | 05 |  |ML|500|78| | | 00810BDF7
                               H-59
                                                                       OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses .
      21 | |L| |C| | |FX1201LF2|1992|05|30| \1992\05\27\ \2.0\ |00811DF1D


      22 | 1392 | 06 | 01 | 11 | 30 | CCSTD01 | | | | 500 | 1 . 0 | | | | 00812FC01


      23 | P | 1992 | 06 | 01 1 11 1 01 1 DFTPP02 | || I I I I  I I  I I MB I ^992 | 06 | 01 1 12 1 10 | FSBLK01 1 P | 0 |
      |  |S|0|R|0|008133228
30|C|108952|3855821|UG/L| | 63 | | A| 75 . 00 | | | | | P | | 84 |D | 2

30 |c| 111444 | |UG/L|BDL| |u| | |u|io| |  | | | | | | | | | | | | |

30|C|95578|385582l|UG/L| | 61 | | A| 75 . 00 | | | | | P| | 82 | D | 1
                                                                  | 008145C61
                                                                 | 00816A3FA
30 | C| 4165600 | 1520963 |UG/L| | 39 .2 | | A| 50 . 0 |


30 |C| 321608 | 15067262 | UG/L | | 40 . 1 | | A| 50 . 0 |
                                                | S | 78 |

                                                | S | 80 |
                                                                  | 0087865AD


                                                                   | 008799002
      30|C|1146652| | | STD | | |AJ20.0| || I I I I I I I I I I I I | I 1 100060 | 00886B4FA
32| | |RT|7.73|
                             00887C42D
                                             ||||| | I | 50488 | 00888E7D2
                                                   | 77500 | 0089019AE
30 |C | 15067262 | | | STD | ||A|20.0|||||||


32| | |RT|9.65| | | | | | | | I  | | 00889F712


30|C|1517222| | |STD| | |A|20.0| | | | 1 || |


32| | |RT|ll.32| | | | | | | |  | | |00891299F


30|C|1520963| | | STD | -| |A|20.0| | || I I I I I I I I I I I | 1 1 29384 | 008924C52


32| | |RT|18.44| | | | | | | |  | | | 008935C6B


30 |C| 1719035 | | | STD | | | A| 20 . 0 | | | | | | | | |


32| | |RT|14.70| | | | | | | |  | | |008958F2D



                               H-60
                                                 | I | 39388 | 008947F23
                                                                       OLM04.2

-------
                                                 Exhibit H -- Appendix A
                                     Format of Records for Specific Uses
30|C|3855821|||STD|||A|20.0||||||||||||||||I|29976|0089SB201


32|||RT|6.37|||||||||||00897C133
                                H-61                               OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
3.0   PESTICIDES  (COL. 1)

3.1   Record Type 10 for Pesticides  (Col 1)

      10 | |  | | | \GC\OLM04.2P\ | TESLAB | | |  |68DOOOOl| INSTP11 | TEST LABS INC. | |000002A30

3.2   Record Type 11 for Pesticides  (Col 1)      . -

      ll|DB-5||0.53||||||||||J0000249FA

3.3   Record Type 20s and 30s for Pesticides (Col 1)  --  Initial  Calibration Mean
      Values (QC Code  'MNC')

      20|||0|MNC||18000|X1201|||||||||61|||000036387


      30|C|319846| |  |AVG|834490| | | | | |R|6|  | | | | |  | | | | | |  | |000048346


      32|||RT|6.92|RTF|6.87|RTT|6.97|||||||000059DCB


      30|C|319857| |  |AVG|272332| | | | | |R|l2| | | | |  | | | | | || I I00006BE34


      32||RT|8.24RTF8.19|RTT8.29||||00007D89C
      30|C|5103742|||AVG|706395||||||R|13||||||||||||||00042FD21


      32 |  | |RT111.87|RTF111.80|RTT111.94| | | | | | |000431AOA


      30|C|8001352| | | |l3757| | | I I I I I I I ! I I I I I I I I I |00044363C


      32|||RT|14.84|RTF|14.77|RTT|14.91||||||l|000455429


      30|C|8001352| | | | 7373 | | || I ! I I I I I 1 I I i ! 1 I I M000466F72


      32|||RT|15.47|RTF|15.40|RTT|15.54||||||2|000478D57


      30|C|8001352| | | |10643| | | | | | | | | | | | | | | | | | | | |00048A976


      32|||RT|16.04|RTF|15.97|RTT|16.11||||||3J00049C748


      30|C|8001352 | | | J17393| | | | | | | | | | | | | | | | || I I |00050E370


      32|||RT|17.23|RTF|17.16|RTT|17.30||||||4|000510136
                                     •H-62                              OLM04.2

-------
                                                       Exhibit  H --  Appendix A
                                           Format of  Records  for Specific Uses
30|C|11096825| | | |48393|
32 | | | RT | 14. 01 | RTF | 13. 94 | RTT | 14. 08 |
                                                | 00120C450
                                             500121E21E
30|C|877098| | |AVG|344348| | | | | |R|lO| || I I I I I I I I I I I 0012202BA


32 | | | RT | 5. 04 | RTF | 4. 99 | RTT | 5. 09 | | | |  | |  | 001231D26


30|C|2051243| |  |AVG| 455516 | | | | | |R| 10 |  | | | | | | | | |


32| | |RT|22.53|RTF|22.43|RTT|22.63|  | |  | | | | 001255B43
                                                       001243E74
3.4   Record Type 20s and 30s for Pesticides (Col 1)  -- Florisil Cartridge Check
      Recovery Values (QC Code 'FLO')

      20 | \FLOP1248A100\0\FLO\ \ 18000 | X1201 | 1991 \ 11 | 14 | 09 | 19 | | | | 11 | | | 00126852C
      21
         | P1248A1
                               2 . 0  00127993B
30|C|319846| |NG| |8.695| |A|lO.OOO| | |


30 | C| 58899 |  |NG| |8.816| |A|lO.OOO| | | |


30|C|76448|  |NG| |8.690| |A|lO.OOO| | | |


30 | C| 959988 | |NG | | 8 . 891 1 | A| 10 . 000 | | |


30|C|6057l|  |NG| |17.498| |A|20.000| | |


30|C|72208|  |NG | | 16. 576 | |A|20.000| | |


30|C|72548|  |NG| |23.206| |A|20.000| | |


30 | C| 50293 |  |NG| |16.551| |A|20.000| | |


30|C|72435|  |NG| | 84 . 668 | | A| 100 . 000 | |


30 | C| 877098 | |NG| |9.882| | A| 10 . 000 | | |


30 |C | 2051243 | |NG| | 20 . 032 | | A| 20 . 000 |
                                         | |S|87.0| | | |


                                         | S | 88 .2 | | | | |


                                         |S|86.9| | | | |


                                         | | S | 88 . 9 | | | |


                                         | |S|87.5| | | |


                                         | | S | 82 . 9 | | | |


                                         | | S 1 116 . 0 | | |


                                         | |S|82.8| | | |


                                         | | | S | 84 . 7 | | |


                                        | | | S | 98 . 8 | | | |


                                            S 100 . 2
                                                           | |00128BFC9


                                                           |00129E581


                                                           J001300B3E


                                                           | |0013131F9


                                                           | | 001325882


                                                           | | 001337EF2


                                                           | | | 00134A61A


                                                           | | 00135CC78


                                                           | | |00136F3EA


                                                           | | 001371A93


                                                           | |    001384337
      30/C/35954//WG//2//A/50.000//////////S/4///////001384338
                                     H-63
                                                                       OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses


3.5   Record Type 20s and 30s for Pesticides (Col  1)  -- GPC Recovery Values  (QC
      Code 'GPC')

      20)\GPC1242A112B\0\GPC\\18000|X1201|1991\11|13|09|19||||11|||00139852C

      21| | | | | |  |G1242A1| |  | I | | | | |2.0| |00140993B

      30|C|58899||NG||9.816||A|lO.OOO||||||||||S|98.2|||||||00141E581

      30|C|76448||NG||9.690||A|lO.OOO||||||||||S|96.9|||||||001420B3E

      30|C|309002||NG||9.89l||A|lO.OOO||||||||||S|98.9|||||||0014331F9

      30|C|60571||NG||18.498||A|20.000||||||||||S|92.5|||||||001445882

      30|C|72208||NG||17.576||A|20.000||||||||||S|87.9|||||||001457EF2

      30|C|50293||NG||17.581||A|20.000)|||||||||S|87.9|||||||00146CC78

3.6   Record Type 20s  and  30s  for  Pesticides   (Col 1)  --   Resolution  Check
      Standard (QC Code  'LPC')

      20||RESC01|0|LPC||18000|X120l|1992\04|20|04|44||||9|||001396901

      21|||||||RESC01|||||||||2.0||001407BA9

      30|C|877098| | |  | | | |  | | |R|98.9| | | | | | || I I  I I I  |0014196FF

      32|||RT|5.06|||||||||||00142A62C

      30|C|5103742| |  | | | |  | | | |R|lOO.O| | | || I I 1  ! 1 I  I I I00143C2DE

      32|||RT|11.88|||||||||||00144D2EE

      30|C|959988| | |  | | | |  | | |R|lOO.O| | | | M I I I  I I I  I I00145EEFC

      32|||RT|12.40|||||||||||00146FEF7

      30|C|72559||||||||||R|92.3||||||||||||||00147193C

      32|||RT|13.10|||||||||||00148292B


      30|C|60571| | | |  | | | |  | |R|100.0| | || I i I I 1 I  I I I  |001494422

      32| |  |RT|13.26|  | | | |  | | | || I 001505422



                                      H-64                              OLM04.2

-------
30|C|1031078| |


32| | |RT|16.03|


30|C|72435| | | |


32| | |RT|17.87|
                         | | |R|100.0| | | | |


                         | I I I |0015280C6

                         |R|98.6| M | | | | |


                              00154AB36
                                                 Exhibit H -- Appendix A
                                     Format of- Records for Specific Uses
                                           0015170C8
                                        001539B17
30|C|53494705| | | || I II I |R|lOO.O| | | | I I I I I I I I I |00155C8F1
      32   RT 18. 15
      30|C|2051243|
      32| | |RT|22.53|
                        00156D8FE
                 ||||||||||||||||||||00157FOAD
                       (0015800AD
3.7   Record Type 20s and 30s for Pesticides  (Col 1)
    •  Mixture  (QC Code 'CLE')
                                               --Performance Evaluation
20 | |PEM01|0|CLE| | 18000 |X1201 | 1992\ 04 | 20 | 06 | 26 | | | | 12 | | | 001592687
               |PEM01|
                       |2.0| |00160389F
      30 | C| 319846 | |NG| | 0.020 | |A|0.020| | | R| 97 . 5 | | | |D | 0 . 0 |


      32| |  |RT|6.94| | | | | |  | | | | |001626C20


      30 | C| 319857 | |NG| | 0 . 022 | |A|0.020| | |R|94.0| | | |D|-10.
      32   RT 8 . 25
                       00164AOED
      30 | C| 58899 | |NG | | 0 . 020 | |A| 0 . 020 | | |R| 93 . 4 | | | | D | 0 . 0 | | |


      32 | | | RT | 8. 03 | | | | | | | | | | |00166D388


      30 | C| 72208 | |NG| |0.117| |A|0.100| | |R|98.5| | | |D|-17.0|
                                                          | 001615CE1
                                                            0016391BC
                                                         | 00165C465
                                                           | 0017048D9
      32) | | RT | 14. 26 |
                    PB | 11 . 0  13 . 5    0017160EC
30 | C| 50293 | |NG||0.201||A|0.200|||R|94.6||||D|-0.5|||||||| | 0017284BB


32| | | RT 1 15 .'44 | | | | | | PB | 2 . 5 1 13 . 5 | | |001739COC


30 | C| 72435 | |NG| |0.54l| |A|0.500| | |R|97.4||||D|-8.2|||||||| | 00174C014


32| | |RT|17.86| | | | | | | | | | |00175D032

                                H-65                              OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
                                                         | 001806222
                                                       0018296ED
      30 |C| 877098 | |NG| | 0 . 038 | | A| 0 . 040 | | |R| 92 . 6 | | | |D| 5 . 0 | |


      32 | |  |RT|5.06| | |  | | | | | |  | |00181714F


      30|C|2051243| |NG| |0.042| |A|0.040| | | | |  | | |D|-5.0| | | | |


      32| |  |RT|22.53| |  | | | | | |  | | |00183A6ED

3.8   Record Type 20s and  30s  for  Pesticides  (Col 1)  --  Initial Calibration
      Multicomponent Standard (QC Code  'CLS')

      20| |AR16600l|0|CLS| | 18000 |X1201 | 1992\  04 | 20 | 06 | 58 | | | | 12 | | | 00184CFD8


      21| |  | |  | |  |AR16600l| | | |  | | | | |2.0| | 00185E4C3


      30|C|12674112| |  | |27928| |A|0.200| | | | | |  | | | | | | |  | | | | | 0018606AO


      32| |  |RT|9.00| | |  | | | | | |  |l|00187169E


      30|C|12674112|.|  | |694l| |A|0.200| | | | I | |  I | | | | | |  | | | | 001883797
32| | |RT|10.80|
                             |2|001894872
30|C|12674112| | | |6398| |A|0.200|
                                                      I 001906971
32 | | | RT | 11 . 14 |
                             3 001917A4B
      30|C|12674112| | | | 9007 | |A|0.200| | | | I I I I I I I I I I I I I I 001929B36


      32| | |RT|9.46| | | | | |  | | | |4|00193AB5F


      30|C|12674112| | | |l082l| |A|0.200| | | | I I I I I I I I M I I II 00194CD18
32   RT 7 . 99 |
                            | 5 | 00195DD48
30 |C| 11096825 | | | | 94264 | | A| 0 . 200 | | | | | | |


32| |  |RT|16.86| | | || I I II |l|001971038


30 | C| 11096825 | | | | 74552 | | A| 0 . 200 | | | | | | |


32| |  |RT|14.23| | | |  | | | |  | |2|001994308


30|C|11096825| | | |32609| |A|0.200| | Ml I I


32 | |  | RT | 18. 14 | | | |  | | | |  | |3|0020175DO

                               H-66
                                                       00196FF3E
                                                      -|00198322C
                                                 I I I I I I I 0020064E5
                                                                       OLM04.2

-------
      30|C|11096825| | | |34595| |A|0.200|
32   RT 19.661
                            |4|00203A8CF
      30|C|11096825| | | |48393| |A|0.200|
      32   RT 14. 01
                       |5|00205DBA2
                                                 Exhibit H -- Appendix A
                                     Format of Records for Specific Uses
                                                 0020297C7
                                                 00204CABD
                                                00206FCDA
      30|C|877098| | | |344348| |A|0.040| | | | | | | | | | |


      32| | |RT|5.05| | | I | | | M I |002070C06


      30|C|2051243| | | |455516| |A|0.040| || I I I II I I I I I I I II I 002082DFO

      32| | |RT|22.54| | | | | | |  | | | |002093DFB

3.9   Record Type 20s and  30s for Pesticides  (Col  1)   --  Initial Calibration
      Single Component -- Individual Standard A (QC Code 'CLM')

      20 | |INDAL13|0|CLM| | 18000 |X1201 | 1992 \ 04 | 20 | 11 | 41 | |  | | 11 | | |00304C753


      21 | | | | | | |INDAL13| | | |  | | | | | 2.0 | |00305DA70


      22||||||||||||||1.0|00306E736


      30 |C | 319846 | | | | 867823 | |A|0.010| | | | || I I I I I

      32| | |RT|6.9l| | | | | | | |  | | |003081789


      30|C|58899| | | |807087| |A|0.010| I I || I I I I I I I
                                             I I I | 003070857
                                               |0030937DC
      32   RT8.0l
                        0031046FD
      30|C|76448| | | | 827300 | |A|0.010| | || I I I I I I I I I I I I I |003116728


      32| | |RT|8.89| | || I I I II I |00312766D


      30|C|959988| | | |605148| |A|0.010| || I I I I I I I I I I I I I I |0031397A9


      32| | |RT|12.38| | || I I I I I I |00314A7AB


      30|C|6057l| || |638273J |A|0.020| | | | I I I I I I I I I I I I I |00315C7CC


      32 | | | RT 1 13. 24 | | | || I I I I I |00316D7CA


      30|C|72208| | | |444069| |A|0.020| | | | | | | | I I I II I M I |00317F7F3

                                      H-67    '
                                                                  OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses


      32|||RT|14.25|||||||||||0031807FD

      30|C|72548| | |  |444235| |A|0.020| | | | | | | | | I I II I I I I I 003192832

      32|||RT|14.62|||||||||||00320383D

      30|C|50293||||401902||A|0.020||||||||||||j|||||00321583D

      32|||RT|15.43|||||||||||003226848

      30|C|72435| | |  |174490| |A|0.100| | | I I I I I I I I I I I I I I |003238870

      32|||RT|17.85|| | | | | | || I|00324988D

      30|C|877098| |  | |373538| |A|0.010| | | I I I I I I I I I I I I I I |00325B9B1

      32|||RT|5.03|||||||||||00326C8D1

      30|C|2051243|  | | |496702| |A|0.020| I I I | I I I I I I | 1 | | | | |00327EAA7

    - 32|||RT|22.52|||||||||| J00328FAA6

3.10  Record  Type 20s and  30s  for Pesticides  (Col  1)  --  Initial Calibration
    '' Single Component  --  Individual  Standard Mix B  (QC Code  'CLM')

    /-. 20 | |INDBL15|0|CLM| |18000|X1201|1992|04|20|12|13| | | 113 | | | 003292199

      21|||||||INDBL15|||||||||2.0||0033034C3

      22|-||  | || | | I'll  | ||1.0|003314189

      30|C|319857| |  | |301500| |A|0.010| | II I I I I I I I I I I I I I |003326275

      32| | |RT|8.25|  | | || I I I I I |0033371A6

      30|C|319868| |  | |946215| |A|0.010| | I I II I II I I I I I I I I I0033492C4

      32| | |RT|9.27|  | | | | I I I I I I00335A1F8
      30|C|7421934| | | |461669 | |A|0.020| | || I I I I I I i I I I I I I (00348EE52

      32| | |RT|15.69| | | | | || | I I |00349FE6F

                                     H-68                    .           OLM04.2

-------
                                                         Exhibit H -- Appendix A
                                             Format of Records for Specific Uses


        30|C|5103719| | | |781319| |A|0.010| | || I I I I I I I I I I I I I |00350204E

        32| | |RT|12.33| MINIMI |003513041

        30|C|5103742| | | |804407| |A|0.010| | | | | | | | | | I I I I I I I I 003525220

        32|||RT|ll.87|||||||||||00353622F

        30|C|877098| | | |373538) |A|0.010| || M I M I I I I II M I |003548353

        32|||RT|5.04|||||||||||00355927E

        30|C|2051243| | | |496702| |A)0.020) | | | I I I I I M I I I I I I |00356B454

        32| | |RT|22.53| | | | | I I I I I I00357C454

  3.11  Record Type 20s and  30s  for Pesticides (Col  1)  --  Instrument Blank  (QC
        Code 'LIB')

|        20||PIBLKX1|1|LIB||18000|X1201|1992\04|21|12|19||ML|lOOO|30|||004663C44

        21|||||||PIBLKX1|||||||||2.0||004674F64

        22) | | | | | | | |  |10000|1.0| | | [004686077

|        30|C|319846| |UG/L|BDL| | | |  |U|0.050\ M M I I I M M  I I I I00469809D

|        30|C|319857| |UG/L|BDL| | | |  |U| 0.050\ | | I | I | | I  | l'|  || I I 00470AOC5

|        30|C|319868| |UG/L|BDL| | | |  |u|0.050\ | | | M M I  M II I I |00471COEF
        30|C|11096825| |UG/L|BDL| | | | |u|1.0\ \ \ \ \ \ \ \ \ \ \ \ \ \ | |00496E7D1

        30|C|877098| |UG/L| |0.040| |A/0.2/| | | I I.I I I I I I I I I I I I I I 0049708D6

        32|||RT|5.06|||||||||||004981803

        30|C|2051243| |UG/L|NRP|0.042| |A/0.2/| | | | I I I I I I I I M I I I I I004993E4D

        32| | |RT|22.54| | | I I I I I I I I005004E58




                                        H-69                               OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
3.12  Record Type 20s and 30s for Pesticides  (Col 1)  -- Method Blank  (QC Code
      •LRB' )

      20 | |PBLK58|l|LRB| | 18000 |X1201 | 1552 | 04 | 21 | 14 | 31 |  |ML|lOOO|30| | (0052620A8

           |S| | |FPBLKOl|l992|04 |17 |  | |  | | |2.0| |0052739F9
      21
      22
                  |10000|1.0| | | J005284BOC
      27| | | | | | | |F|1991|11|14|09|19|P1248A1|N| | | | | | | | |005308E9E

      30 | C| 319846 | |UG/L|BDL| | | | |U| 0 . 050 | | | | | | | | | | | | | | | |00531AEC2


      30 | C| 319857 | |UG/L|BDL| | | | |U|0.050| 11111111111111 I 00532CEE8
      30|C|11096825| |UG/L|BDL| | | | |U|l.O| || I I I I I I I I I I I I I 00558082B

     : 30 | C| 877098 | |UG/L| | 0 . 19 | | A| 0. 2\ \ \ \ \ \ \ \ \ \ S | 94 | | | | | |  | 005592E19

     " 32| | |RT|5.06| | | | | | | |  | | |005603D46

      30 | C| 2051243 | |UG/L| |  0 . 21 | | A| 0.2 \ \ \ \ \ \ \ \ \ \ B \ 100 | | | |  | | | 005616468

      32| | |RT|22.53| | | | | | |  | | | | 005627468

3.13  Record Type 20s and 30s for Pesticides (Col 1)  -- Matrix  Spike Sample  (QC
      Code ' LF1 ' )

      20| |X1201MS|1|LF1| | 18000 JX1201 | 1992 | 04 | 21 | 16 | 08 | |ML|300|30| | | 00563A08B


      21 | | | |S| | | 488920 | 1992 | 04 | 17 | |l992|04|l5| |2.0| |00564BFE5
      22
      23
                  | 5000 1 1 . 0 | | | | 00565D2D6
            | | | | | | | | | | | | | | |MB|1992|04|21|14|31|FPBLK01|P|0| | | S | 5 | R| 5 | 005660949


      27 | | | | | | | |F|1991|11|14|09|19|P1248A1|N| | | | | | | | |00567285F


      30 | C| 319846 | |UG/L|BDL| | | | \\J\0.050\ 1 | | | | | | | || I I I I I 005684889
                                     H-70
                                                                       OLM04.2

-------
                                                      Exhibit H -- Appendix A
                                          Format of-Records for Specific Uses
30|C|58899| |UG/L| | 0 . 82 | |A|l.667| | | | |P|49| |D|l


32| | |RT|8.01| | | |  | | | | |  | J00572B839


30|C|76448| |UG/L| |0.69| |A|l.667| | | | | P | 41 | |D|2|
                                                        |Pl2.3| | | 00571A918
                                                       |P|6.2| | | 00573D7A3
32| | |RT|8.90|
                             00574E6D6
      30|C|50293| |UG/L| |l.5| |A|l.667| | | | |P|90| |D|6| | | | |P|0.0| | | 005875913
32| |  |RT|15.43|
                             |00588691E
30 | C| 72435 | |UG/L|BDL| | | | |u| 0. 50 \ \ \ \ \ \ \


30|C|53494705| |UG/L | LLS | 0 . 094 | J | | | | | | |
                                                    |00589877C


                                                    |P|6.4   | 00590AD22
32| | |RT|18.14|
                             |00591BD2E
30|C|7421934| |UG/L | LLS | 0 . 091 1 j|
                                             I I I I I I IPI2-2I I I 00592E1B6
32   |RT|15.68
                             | 00593F1D2
      30 |C| 5103719 | |UG/L|BDL| | | | |u| 0. 050 \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ | 0059412C9
30 | C| 877098 | |UG/L| | 0. 20\ \ A| 0. 20 \ \ \ \ \


32 | | |RT|5.03| | | | | | | | | | |0060568AC


30 | C| 2051243 | |UG/L| | 0. 20\ \ A| 0.20\ \ \ \ \ \ \ \ \ \ S | 100
                                             \ \ S \ 100\ \ \ \ \ \ | 00604598C
                                                            006068FOA
 32| | |RT|22.52|
                              006079F09
3.14  Record Type 20s and 30s for Pesticides (Col 1)  -- Matrix Spike Duplicate
      Sample (QC Code 'LF2')

      20| |X1201MSD|l|LF2| | 18000 JX1201 | 1592 | 04 | 21 | 16 | 4i | |ML|300|30| | |00608CBB4


      2l| | | |S| | |488921|1992|04|17| |1992|04|15| |2.0| | 00609EBOF

                                     H-71                              OLM04.2

-------
Exhibit K -- Appendix A
Format of Records for Specific Uses
      22| |  | | | | | | | |5000|l.0| | | |00610FEOO


      23||||||||||||||||||MB|1552|04|21|14|31|FPBLK01|P|0|||S|5|R|5|006113473


      27 | |  | | | | | |F|1S91|11|14|09|19|P1248A1|N| | | | | | | | |006125389


      30 |C| 319846 | |UG/L|BDL| | | | \\J\0.050\ | | | | | || | | | | | | | | 0061373B3
      30|C|76448||UG/L||0.70||A|l.667|||||P||42|D|2|||||P|6.l||[0061802BC


      32|||RT|8.9l|||||||||||0061911FO
.30 | C| 50293 | |UG/L| |l.6| |A|l.667| ||||


:32| | | RT 1 15. 45 | | | | | | | | | | |00633944F


?30|C|72435| |UG/L|BDL| | | | |U| 0. 50 | | | |


K.30|C| 53494705 | | UG/LJLLS | 0 . 098 | J| | | |


 32| | |RT|18.16| | | | | | | | | | |00636E928


 30 |C| 7421934 | JUG/L | LLS | 0 . 096 | J| ||||


 32 | | | RT | 15. 70 | | | | | | | | | | (006381DD7


 30|C|5103719| |UG/L|BDL| | | | |U| 0.050\
                                                                006328438
                                                      00634B2AD
                                                     |P|l2.2| | |00635D91A
                                                            |006370DCC
                                                         006393ECE
      30 |C | 877098 | |UG/L| | 0. 19\ \ A| 0. 20 \ \


      32| | |RT|5.04| | | | | | | | | | |0065094B5


      30 |C | 2051243 | |UG/L| |0.18||A|0.20|
                                     I I I I IsI97I I I I I I |00649858A
                                                      00651BB13
 32 | | | RT | 22 . 53 |
                              | 00652CB13
                                     -H-72
                                                                        OLM04.2

-------
                                                      Exhibit H  -- Appendix A
                                          Format of Records  for  Specific Uses
3.15  Record Type 20s and 30s for Pesticides (Col 1)  -- Continuing Performance
      Check -- Individual Standard Mix A (QC Code 'CLE')

      20 | | INDAM72 | 0 | CLE |  | 18000 | X1201 | 1992 \ 04 | 21 | 22 | 12 | | | | 11 |  | |006882FOF

      21 1 | | | |  | | INDAM72 ||  I I I  I II I 2 • ° I  I 00689423C

      22| | | | |  | | |  | | | | | |4.0|006904FOF

      30 |C| 319846 | |NG| | 0 . 039 | | A| 0 . 040 |  | | | | |  | |D | 2 . 5 | |  | |  | | | |  | 006917382

      32 | | |RT|6.94| || I I I  I I I  I I 0069282C1

      30|C|58899| |NG| | 0 .  035 | |A|0.040| |  | |

      32 | | | RT | 8. 03 | |  | | | |  | | | | |00694B666
                                     |D|l2.5|
                                                           00693A743
30|C|72435| |NG||0.438||A|0.400||||||||D|-9.5||

32| | |RT|17.86| | | | | I I I I I |00708234C

30 | C| 877098 | |NG| |0.037| |A|
                                              |D|23.8|
32| | |RT|5.05| | | | | | |  | | | |00710570A

30|C|2051243| |NG| |0.06l| |A|0.080| |

32| | |RT|22.52| | | | I I  I I I I |007128D17
                                                          | 00707132E
                                                          | 0070947DE
                                                             007117D18
3.16  Record Type 20s and 30s for Pesticides (Col 1)  -- Continuing Performance
      Check -- Individual Standard Mix B (QC Code 'CLE1)

   -  20 | |INDBM78|0|CLE| | 18000 |X1201 | 1392 | 04 | 21 | 22 | 51 |  | | | 13 | | J00713B40B

      21 1 | | | |  | | INDBM78 |  | | | |  | I I I 2 • ° I  I 00714C749

      22|||||||||||i||4.0|00715D41C

      30 | C| 319857 | |NG| |  0 . 036 | | A| 0 . 040 |  | | | | |  | | D 1 10 . 0 |.| |  | | | |  | | 00716F95A

      32| | |RT|8.26|  | | Ml I I I  I | 00-717088C
                                     H-73
                                                                       OLM04.2

-------
Exhibit K -- Appendix A
Format of Records for Specific Uses
 30 |C| 319868 | |NG| | 0 . 038 | | A| 0 . 040 | |


 32| | |RT|9.28| | | | | | | |  | | [007193C35


 30 |C| 309002 | |NG| |0.038| |A|0.040| |


 32| | |RT|9.83| | | | | | | |  | | | 007216FAC
                                            |D | 5 ..0 |
                                            |D|5.0|
                                                           | 007182DOO
                                                            007206076
 30|C|5103742| |NG| |0.038| |A|0.040| |


 32| | |RT|ll.88| MINIMI |007371E78


 30 | C| 877098 | |NG| |0.042| |A|0.040|||


 32| | |RT|5.06| || || || | || |00739529B


„. 30 |C| 2051243 | |NG| |0.067| |A|0.080| |
                                             |D|5.0|
                                              D 16 . 2
                                                            |007360E68
                                                            |00738436E
                                                             |0074078B5
     . 32| | |RT|22.54|
                              I 0074188CO
3.17  Record Type 20s and 30s for Pesticides  (Col 1) -- Continuing Performance
    '"Check -- Performance Evaluation Mixture  (QC Code  'CLE')

      20| |PEM90|0 |CLE| | 18000 |X1201 | 1992\ 04 | 23 | 13 | 47 | | |  | 12 | | | 00777E850


      2l||||||| PEM90 | | | |  || || | 2 . 0 || 00778FA7A


                                                                    007792031
                                                                  | 00781551E
                                                                 |OO783891F
                                                                |007880CB5
      30 |C| 319846 | |NG| |0.025| |A|0.020| | |j?/99.4/| | | | |D|-25.0|


      32| | |RT|6.93| | | | | | | | | | |007802F65


      30 |C| 319857 | |NG| |0.025| |A|0.020| | |R| 99.4 | | | |D| -25.0J | |


      32| | |RT|8.25| M I I I I I M | 00782644F


      30 | C| 58899 | |NG| |0.022| |A|0.020| | |R|98.7| | | |DJ-10.0| | | |


      32| | |RT|8.03| | | | | | | M I |007849842


      '30|C|72208| |NG| |0.108| |A.|0.100| | |R|98.9| | | |D|-8.0| | | ||
      32 | | | RT | 14. 27 |
                         | PB | 20 . 0 | 24 . 0 | | | 0078924C6
                                     H-74 '
                                                                        OLM04.2

-------
                                                         Exhibit H -- Appendix A
                                             Format of Records for Specific Uses


        30|C|50293||NG||0.201||A|0.200|||R|95.9||||D|-0.5|||||||||007904895

        32|||RT|15.45||||||PB|4.0|24.0|||007915FE1

        30|C|72435| |NG| |0.518| |A|0.500| | |R|97.9| | | |D|-3.6| | | | | | | | |0079283EC

        32| |  |RT|17.86| | | |  | | | | | | |00793940A

        30|C|877098||NG||0.044||A|0.040|||R|98.7||||D|-10.0|||||||||007982735

        32|||RT|5.05|||||||||||007993661

        30|C|2051243| |NG| |0.04l| |A|0.040| | | | | | | |D|-2.5| | | | | | | | |008005COO

        32| | |RT|22.53| | | | | | | | | | |008016COO

  3.18  Record Type 20s and 30s for Pesticides  (Col 1)  -- Field Sample  chosen  for
        MS/MSD (QC Code 'LSD')

J        20||X120l|l|LSD||18000|X1201|1SS2|04|23|15|42||ML|400|34|||0080296B4

|        21||||S|||FX1201|1932|04|17||l332|04|l5||2.0||00803B55E

        22| | | | | | | | | |500011.0 | | | |00804C84F

|        23||||||||||||||||||MB|1392|04|2l|l4|31|FPBLKOl|P|2| | I I I I I00805FBAE

|        27| | | | M | |F|19Sl|11|14|09|19|P1248A1|N| | | I I II I |008061AC4

|        30|C|319846| |UG/L|BDL| | | | |u|0.050\ | | | || I I I I I I I I I I008073AEB

|   .     30|C|319857||UG/L|BDL|||||U|0.050||||||||||||||||008085B14

|        30|C|319868| |UG/L|BDL| | | | |u|0.050\ | | | I I I I I I I I I I I I008097B3F
         30|C|72559||UG/L|LLS|0.014|JP||||||||||||||||P|27.3|||008165DD3

         32|||RT|13.03|||||||||||008176DC4

         30|C|72208| |UG/L|BDL| | | | \U\0.10\ | | | I | I I I I I | | | | |008188C04

         30|C|33213659||UG/L|BDL|||||u|0.10\|||||||||||||||00819ACEF

                                        H-75                              OLM04.2

-------
Exhibit K -- Appendix A
Format of Records for Specific Uses
30 |C| 877098 | |UG/L| |0.10||A|0.20|||


32 | | | RT | 5. 04 |  | | | || | | | | |00847F5F6


30 | C| 2051243 |  |UG/L| | 0.10\ |A|0.20| |


32| | |RT|22.53| | | | | | | | | | |008492C7E
                                                              |00846E6CB
                                                |S|53| | |P|7.7| | |008481C7E
                                     H-76
                                                                        OLM04.2

-------
                                                       Exhibit H -- Appendix A
                                           Format of Records for Specific Uses
4.0   PESTICIDES (COL. 2)

4.1   Record Type 10 for Pesticides  (Col 2)

      10 | | | | | |GC| OLM04.2P\ | TESLAB | | |  | 68D000011INSTP2 | | TEST LABS INC. | | 000002A31

4.2   Record Type 11 for Pesticides  (Col 2)

      11|DB-608| I 0.53 I | | | | |  | | | | |000024AD7

4.3   Record Type 20s and 30s for Pesticides  (Col 2)  --  Initial Calibration Mean
      Values (QC Code  'MNC')

      20|||0|MNC||18000|X1201|||||||||62|||000036465


      30|C|319846|||AVG|1137869||||||R|16||||||||||||||0000485E4


      32| | |RT|4.56RTF|4.51 | RTT|4.611 | | | | | |00005A047
      30|C|12674112| | | |35883| | | | || I I I I I I I I I I I II I I00054566A

      32|||RT|5.69|RTF|5.62|RTT|5.76||||||l|0005571C3


      30|C|12674112| | | 1161411 | | | II II I I I I I I I I I I I ||000568ECC


      32|||RT|4.70|RTF|4.63|RTT|4.77|||||I2I00057AA13


      30|C|12674112| | | 118842 | | | I I I I I I I I I I I I I I I I I |00058C730

      32|||RT|6.51|RTF|6.44|RTT|6.58||||||3|00059E285


      30|C|12674112| | | |l5946| | | | I I I I I II I I I I I I I I I |00060FFAE

      32| | |RT|7.30|RTF|7.23 |RTT|7.37||||||4|000611AEO


      30|C|12674112| | | |l6309| | | | | | | | | I I I I I I I I I I I I0006237EF

      32|||RT|8.01|RTF|7.94|RTT|8.08||||||5|00063533C


      30|C|11104282| | | |9686| | | | | | I I I I I I I I I 1 I I I I I000646F7E

      32|||RT|3.89|RTF|3.82|RTT|3.96||||||l|000658AB9
                                      H-77                              OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
      30|C|11104282| | | |3312 | | I 1 | I || I I  I I I I I I 1 I I | |00066A6BF


      32|||RT|3.49|RTF|3.42|RTT|3.56||||||2|00067C1EF
4:4
      30|C|877098| | |AVG|559374| | | |  |.|R|2l| 1  | | | | | | | | I I I (001242054


      32 | | | RT | 2. 87 | RTF | 2. 82 | RTT | 2.  92 | | | | | |  | 001253A9F


      30 |C| 2051243 | |  |AVG| 594715 | | |  | | | R| 16 |  | | | | | | | | | | | | |001265C02
32| | |RT|16.34 | RTF | 16. 24 | RTT | 16. 44 |
                                             | 0012778FF
Record  Type  20s  and 30s  for Pesticides  (Col  2)  -- Resolution  Check
Standard (QC Code 'LPC')

20 | |RESC01|0|LPC| | 18000 |X1201 | 1992 | 04 | 20 | 04 | 44 | | | |9| |  | 0014186BD
 211 | | | | | |RESC01|


•30|C|877098 Mill


 32| | |RT|2.89| | I I

 30|C|959988| | | |-|


 32|||RT|8.71||||


 30|C|5103742||||


 32| | |RT|8.89| | | |


 30|C|72559| | | II I


 32| | |RT|9.35| | | |


 30|C|60571||| N I


 32| ||RT|9.611| | |


 30|C|1031078| | | |


 32M IRT|13.17| I I
                           | | | |2.0| [001429965


                           |R|100.0||||||||||||||00143B560


                           | | | 00144C495
                    | R | 97 . 0 |
                             00146EF1C
                          | | | |001481B13


                          |R|99.0MNNMII

                          | | | | 001504499

                          |R|IOO. OMNIUM

                          | | | |001526EC2


                          M|R|96.3|||||NI

                          | | | | |001549AB1

                                     •H-78
                                               00145DFEA
                                                 |001470BCE
                                      | | 001493566
                                        |001515F90
                                         |001538AB1
                                                                       OLM04.2

-------
30|C|72435| | | |


32| | |RT|13.35|
                   |R|lOO.O|
                                                       Exhibit H -- Appendix A
                                           Format of -Records for Specific Uses
                                               00155B5AA
                              00156C5AA
      30|C|53494705| | | | | | | | | |R|lOO.O| | | | | | | | | | | | | | 00157E365


      32| | |RT|14.06| | | | | | | M | |00158F36E


      30|C|2051243|||||||||||||||||||||||||001590B1D


      32 | | |RT|16.35| | | | | | | | | | |001601B2A

4.5   Record Type 20s and 30s for Pesticides (Col 2) -- Performance Evaluation
      Mixture (QC Code 'CLE')

      20| |PEM01|0|CLE| | 18000 JX1201 | 1992 | 04 | 20 | 06 | 26 | | |  | 12 | | | 001614104
21
               PEM01
                             2.0  00162531C
30 | C| 319846 | |NG| |0.02l| |A|0.020| | |R|99.l| | | |D|-5.0|


32| |  |RT|4.59| | |  ||  I I I I I |001648725


30 | C| 319857 | |NG| |0.02l| |A|0.020| | |R|94.9| | | |D|-5.0|


32 | |  | RT | 7. 31 1 | |  | |  | | | | | |00166BB15


30 | C| 58899 | |NG|  |0.020| |A|0.020| ||R|97.8||||D|0.0|||


32| |  |RT|5.47| MINIMI I 00168EDC9


30 | C| 72208 | |NG |  | 0.103 | | A | 0.100 | | |R | 95 . 6 || | |D | -3 . 0 | |


32| |  |RT|10.09| || II |PB|11.0|12.0| | | 0017379E1


30|C|50293| |NG |  | 0.182 | |A| 0 . 200 | | | R| 96 . 3 | | | | D | 9 . 0 | | |


32| |  |RT|11.68| |  | | | |PB|1.0|12.0| | | 00175B476


30 | C| 72435 | |NG|| 0 . 501 | | A| 0 . 500 | | | R| 94 . 8 | | | | D | -0 . 2 | |


32 | |  | RT | 13. 34 ||  | | | | | | | | |00177E85D


30|C|877098| |NG| |0.037| |A|0.040| | |R|95.8| | | |D|7.5| |


32| |  |RT|2.89|| | | | M I I I |001838971

                                H-79
                                                                 0016377E7
                                                                 00165ABF2
                                                               00167DE8D
                                                                |0017261EB
                                                               | 00174 9D4D
                                                                 00176D85E
                                                                 001827A3C
                                                                        OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
      30 |C| 2051243 | |NG| | 0 . 039 | | A| 0 . 040 | |


      32| |  |RT|16.34| | | | | |  | |  | | | 00185BEAA
4.6
4.7
                                             |D| 2 . 5
      Record Type  20s  and 30s  for Pesticides  (Col 2)
      Multicomponent Standard (QC Code ' CLS ' )

      20| |AR122102|0|CLS| |'18000 |X1201 1 1992 | 04 | 20 | 07 | 31 1
                                                            00184AE9E
                                                        -- Initial Calibration
                                                              J002128054
      21
              |AR122102|
                                | 2.0 | | 002139525
      30 | C| 11104282 |


      32| |  |RT|3.89| |

      30 |C| 11104282 |


      32| |  |RT|3.49| |

      30|C|11104282|

      32 |  RT|5.67| |
                      | 9686 | | A| 0 .400 |
                            1 00215C641
                      | 3312 | |A| 0.400 |
                            |2|00217F71E
                      |3087| |A|0.400|
                              002192825
      30 | C| 11104282 | | | |2207| |A|0.400| | | |


      32| |  |RT|7.05| | | | | | | | | |4|002215916

      30|C|11104282| | | |2134| |A|0.400| | | |
      32 | | |RT| 8. 02 |
                            |5|0022389FB
      30 | C| 877098 | | | | 559374 | | A| 0 . 040 | | | |


      32 | |  |RT|2.86| | | |  | | | | | | |00225BA76


      30|C|2051243| | | |594715| |A|0.040| | |


      32| |  |RT|16.33| | |  | | | | | | | (00227EC66
                                                      | 00214B62E
                                                      00216E70E
                                                  I I I I I 002181808
                                                       0022048FE
                                            I I I I I I I I I I I 0022279EE
                                                       00224AB44
                                                       I 00226DC65
      Record  Type 20s and  30s  for Pesticides  (Col 2)  --  Initial Calibration
      Single  Component --  Individual Standard A (QC Code  'CLM')

      20 | |INDAM16|0|CLM| | 18000 JX1201 | 1992 | 04 | 20 | 12 | 46 | | | | 11 | | | 00362366C

      21 | | | | | | |INDAM16| | | | | | || | 2.0 | | 003634997
                                     H-80
                                                                       OLM04.2

-------
                                                 Exhibit H --  Appendix A
                                     Format of Records for Specific Uses
22
                4.000364566A
30 | C| 319846 | | | | 1128992 | |A|0.040| | | R| 94 . 7 |


32| |  |RT|4.56| | | |  | | | | | | |0036687A3


30|C|58899| | | |l040769| |A|0.040| | |R|96.3| |
32| | |RT|5.44|
                       00368B815
30|C|76448| | | |l070468| |A|0.040| | |R|96.8|
                                                I I I I I |003657868
                                                    I 00367A8DC
                                                     00369D943
32| | |RT|5.81| | | | | | | M | |00370E873


30|C|959988| | | | 834736 | |A|0.040| | |R|95.7| || | || | I I I I I I  I 0037109C3


32| | |RT|8.68| | | || I I I  I I |003721905
30|C|60571| | | |842038| |A|0.080| | |R|95.7|


32| | |RT|9.58| | | | | | | | | | [003744874


30 | C| 72208 | | | | 630027 | | A| 0 . 080 | | | R| 98 . 1 |
                                              I II M I (003733932
                                                    003756886
32   RT 10. 07
                        003767882
30|C|72548| | | | 599176 | |A|0.080| | |R|92.6| | || I I I I I I I I I |0037798EO
 32| | |RT|11.29|
                        00378A8E1
 30|C|.50293| | | |572605| |A|0.080| | |R|99.2|
 32  | | RT | 11. 66 |
                        00380D91A
'30|C|72435| | | | 237.099 | |A|0.400| | |RI 95 • 5 I I


 32| | |RT|13.33| MINIMI | 00382093E


 30 | C| 877098 | | | | 558724 | | A| 0 . 040 | | |R| 96 . 3 |


 32| | |RT|2.87| || I I I I I I I |0038439B8


 30 |C| 2051243 | | | | 595863 | |A|0.080| I I I I I I I I


 32| | |RT|16.33| | | | | | | | | | |003866BB1


                                H-81
                                                  M I 00379C90E
                                                   I |00381F94A
                                                      003832A85
                                             I I I I |003855BBO
                                                                  OLM04.2

-------
  Exhibit H -- Appendix A
 .Format of Records for Specific Uses


  4.8   Record Type 20s  and 30s for Pesticides  (Col 2)  --  Instrument Blank (QC
        Code  'LIB')

        20||PIBLKB2|l|LIB||18000|X1201|1592|04|21|12|19||ML|lOOO|30||[004708102

        21| | |  | | | |PIBLKB2| | |  | | | | | | 2 . 0| |004719422

        22 | | |  | | | | |  | 11000011.0| | | |00472A535

        30|C|319846||UG/L|BDL||||\V\0.050\|||||||||||||||00473C55B

        30|C|319857||UG/L|BDL|||||U|0.050||||||||||||||||00474E583

        30|C|319868||UG/L|BDL|||||U|0.050||||||||||||||||0047505AD
I        30|c|no9769i||UG/L|BDL|11||u|i.o|11 M I M M M 11 I|oo4990B8E

|        30|C|11096825| |UG/L|BDL| | | | |U|l.O| M M M I I M I I II I005002C8F

|       ?~30|C|877098||UG/L||0.040\\AJO.2\||||||||||||||M M I005014D94

        32| | |RT|2.89| M I M I I I I |005025CC9

|        30|C|2051243| |UG/L|NRP|0.Oil| |A/0.2/| | II I M!| | I | I I | |  | I |005038305


        32| ||RT|16.35| | | || I I II I I 005049312

  4.9   Record Type  20s and 30s for Pesticides  (Col 2) -- Method Blank  (QC Code
        1LRB')

|        20||PBLK58|l|LRB||18000|X1201|1992|04|21|14|31||ML|lOOO|30|||005306672

|        21M | |S| | |FPBLK01|1992|04|17| || | ||2.0| |005317FC3

        22||||||||||10000|1.0||||0053290D6


        23|||||||||||||||||||||||||P|0|||||||00533B552

|        27| M II I I |F|1991|11|14|09|19|P1248A1|N| M M I I I |00534D468

        30|C|319846||UG/L|BDL|||||u|0.050| M M M IM M M I|00535F48C



                                       H-82                              OLM04.2

-------
                                                       Exhibit H --  Appendix A
                                           Format of Records for Specific Uses
      30|C|319857| |UG/L|BDL| | | | |u|0.050| | | | | | | | | I I II I I I0053614B2


      30|C|319868| |UG/L|BDL| | | | |u|0.050| | | | | | | | | I I I I I I |0053734DA


      30|C|58899||UG/L|BDL|||||U|0.050|||||||||||| MI[005385434
      30|C|11096825||UG/LJBDL|||||u|1.0||||||||||||||||005624DF5

      30|C|877098||UG/L||0.18||A|0.20||||||||||S|89|||||||0056373E6


      32| | |RT|2.89| | | | |  || I I I I 00564831B

      30|C|2051243| |UG/L| |0 .211 |A|0.20| | | | | | | | | |S|lOO| | | | | | |00565AA3D

      32 | | | RT | 16 . 35 | | | |  | | | ]• | | | 00566BA4A

4.10  Record'Type 20s and  30s  for Pesticides -- Matrix Spike  Sample (QC Code
      •LF11)

      20||X1201MS|l|LFl||18000|X1201|1992\04|21|16|08||ML|300|30|||00567E66D

      2l||||S|||488920|1552|04|17||l952|04|l5||2.0||0056805C7

      22| | |  | | | |  | | |5000|1.0| | | |0056918B8

      23||||||||||||||||||MB|1992|04|21|14|31|FPBLK01|P|0|||S|1|R|1|005704FOF

      27| | |  | | | |  |F|1991|11|14|09|19|P1248A1|N| | |  | | || I I005716E25


      30|C|319846 | |UG/L|BDL| | | | |U|0.050| | | | | | ||  I I I I I I I |005728E4F

      30|C|319857| |UG/L|BDL| | | | |u|0.050| | | | | || II I I I I I I |00573AE7B


      30|C|319868| |UG/L|BDL| | | | \U\0.050\ || | I I I I I I I I I I I I00574CEA9

      30|C|58899| |UG/L| |0.73 | |A|0.833| | | | |P|88| |D|0| | | | |P112.3| | |00575F7AO


      32|||RT|5.43|||||||||||0057606CE

      30|C|76448||UG/L||0.65||A|0.833|||||P|78||D|1|||||P|6.2|||005772EF6


      32| | |RT|5.81| | M I I I I I I |005783E26

                                     H-83                              OLM04.2

-------
Exhibit H -- Appendix A
Format of Records  for Specific Uses
      30|C|50293| |UG/L| | 1 . 5 | |A|l.667| | | | | P | 90 | |D | 6 | | | | | P | 0 . 0 | | |00591C120


      32 | | |RT|11.66| | | | | | | | | | |00592D12C


      30|C|72435| |UG/L|BDL| | | | |U| 0. 50\ \ \ \ \ \ \ \ \ \ \ \ \ \ \ |00593EF8A


      30 |C| 53494705 | |UG/L| LLS | 0 . 10 | j| ||||||||


      32| | |RT|14.03| | | | |'| | | | | (005952430


      30 |C| 7421934 | (UG/L | LLS | 0 . 089 | j| | | | || I I I


      32| | |RT|12.34| | | | | | | | | | (0059758C7
                                                             005941434
                                                 I I I I p I 2 • 2 I I I 0059648C9
| S | 97 |
                                                          (00608C07E
4.11
      330 |C| 877098 | |UG/L| | 0. 19\ \ A\ 0. 20 \ \ \ \ | |


      ^32 | | |RT|2.86| | | | | | | | | | |00609CFBO


      30 |C| 2051243 | |UG/L| | 0. 20\ \ A\ 0.20\ \ \ \ \ \ \ \ \ \ S \ 100 \ \ \ \ \ \ | 00610F61C


      32| | |RT|16.33| | | | | | | | | | |00611061D

      Record Type  20s and 30s for Pesticides -- Matrix Spike  Duplicate  Sample
      (QC  Code  'LF2')

      20 | |X1201MSD|1|LF2| | 18000 (X1201 | 1952 | 04 | 21 | 16 | 41 | |ML|300|30| | | 0061232C8


      21| | | |S| | |488921|1S92|04|17| |1952|04|15| |2.0| (006135223


      22| | | | | | | | | |5000|l.0| | | |006146514


      23 | | | | | | | | | | | | | | | | | | MB | 1992 | 04 | 21 1 14 | 31 1 FPBLK01 |P|0| | | S'| 1 1 R 1 1 1 006159B6B


      27 | | | | | | | |F|1391|11|14|09|19|P1248A1|N| | | | | | | | | 00616BA81


      30 |C|319846 | |UG/L|BDL| | | | |U| 0. 050\ | | | | | | | | | | | | | | | 00617DAAB
                                     H-84
                                                                        OLM04.2

-------
            Exhibit H -- Appendix A
Format of Records for Specific Uses
30 | C| 76448 | |UG/L| |0.66| |A|0.833| |


32| |  |RT|5.82| | | |  | | | | | |  |006238A90
30 | C| 50293 | |UG/L| |l.6||A|


32| |  |RT|ll.67| | |  | | | | | | | |006371D9D


30|C|72435| |UG/L|BDL| | | |  |U|0.50| | |


30 | C| 53494705 | |UG/L| LLS | 0 . 11 | j| | | |


32| |  |RT|14.05| | |  | | | | | | | |006407171


30|C|7421934| | UG/L | LLS | 0 . 092 | J | | | |
32| | |RT|12.36|
                        00642A607
30|C|5103719| |UG/L|BDL| | | | |U| O.OSO\\


30|C|5103742| |UG/L|BDL| | | | \\S\0.050\ |
 30 | C| 877098 | |UG/L| | 0. 19\ \ A| 0.20 | | |


 32| | |RT|2.87| | | | | | | | | | | 006541CEC


 30 |C| 2051243 | |UG/L| | 0. 19\ \ A\ 0. 20\ \


 32| | |RT|16.35| | | || I I I I I |006565365
                                      |79|D|l|
                     006227B5F
                                          I I
         ||||P|0.0|| |006360D90




         | | 006383BFB


         | P | 12 . 2 | | | 0063 96169




         |P|4.3| | | 006419607




         | | | | |00643C6FE


         I I I I I 00644E7F1
                                         S  96
                                       \ \ \ \ S | 94 |
                006530DB9
                 006554358
                                H-85
                             OLM04.2

-------
Exhibit H  -- Appendix A
Format of  Records  for Specific Uses
4.12  Record  Type 20s  and  30s  for  Pesticides  (Col  2)  --  72 Hours  Aroclor
      Standard  (QC Code  'CLE')
20 | |AR122102|0|CLE| 1 18000 | X1201 1 1532 1 07. | 20 | 06 | 23 | | | |?| |


21| | | | | | |VAR1221V2| | || II I I I2-0| |002139525


30|C|11104282| |DB/L| \200\ | | | | | | | M | | I I I I I I I I |00214B62E


32| | |RT|3.89| | | || I I I I 1 1 1 00215C641


30|C|11104282| | C7G/L| |200| | | | | | | | | | | | | | | | | | | | |00216E70E


32| | |RT|3.49| | | | | | | | | |2|00217F71E


30|C|lll04282||aG/L||200| | | || || | IIIIIIIIIIH |002181808

32| | |RT|5.67| | | | | | | | | |3|002192825


30|Cllll04282| |C7G/L| |200| | | | | | | | | | | | | | | | | |
                                                              | 002128054
32 | | | RT | 7. 05 |
                            | 4 | 002215916
30 | C| 11104282 | \UG/L\ \200\
    | RT | 8. 02 |
                            |5|0022389FB
30 | C| 877098 | \UG/L\ \ 20 \ \ A| 0 . 040 | | |


32 | |  |RT | 2. 86) | | | | | | | | | |00225BA76


30|C|2051243| |DG/L| |20| |A|0.040| |
                                                    0022048FE
                                                   |0022279EE
                                                      | 00224AB44
                                                       |00226DC65
32 | | | RT | 16 . 33 |
                              | 00227EC66
4.13  Record Type 20s and 30s for Pesticides (Col 2) -- Field Sample chosen for
      MS/MSD  (QC Code  'LSD')

      20 | |X120l|l|LSD| | 18000 | X1201 | 1552 | 04 | 23 | 15 | 42 | |ML|400|34| | | 008061655
21
22)
23
                | FX1201 | 1992 | 04 | 17 | | 1992 | 04 | 15 | | 2 . 0 | | 0080734FF
                  |5000|l.0  |   0080847FO
                    MB 1552 04 21 14 31 FPBLK01  P  2
                                                                 008097B4F
                               H-86
                                                                        OLM04.2

-------
                                                Exhibit H -- Appendix A
                                    Format of Records for Specific Uses
27| | | | | | | |F|1991|11|14|09|19|P1248A1|N|


30|C|319846| |UG/L|BDL| | | | |u| O.OSO\ | | | |  |


30 | C| 319857 | |UG/LJBDL| | | | |U|O.OSO| | | | |  |
30|C|6057l| |UG/L|BDL| ||||U|0.


30|C|72559| |UG/L |LLS | 0 . Oil | JP |
32 | | |RT|9.27|
                       00821EC9B
30|C|72208| |UG/L|BDL| | | | |u| 0.10\ || I I I


30 |C| 33213659 | |UG/L|BDL| | | | |U| 0. 10 \ \ \


30|C|72548| |UG/L|BDL| | | | |U| 0.10\ | | | | |
30|C|11096825| |UG/L|BDL| | | | |U|l.O


30 |C| 877098 | |UG/L| | 0 . 26 | | A| 0.20\ \


32| |  |RT|2.86|  | | | |  | || I I  |00851720A


30 | C| 2051243 |  |UG/L|  | 0 . 28 | | A| 0. 20\
                                              | 008109A65


                                              | I |00811BA8C


                                                 00812DAB5
                                            || | 00819B90E


                                            |P|27.3| | | 00820DD67




                                            I I I 008220ADB


                                            | | | | | j 008232BC6


                                            | | | 008244A21
                                         IIIIIIMI008493D1C


                                         |S|130|||||||0085062D8
                                          |S|140|.
                                                      I0085298AO
 32| | |RT|16.34|
                        00853A8AC
                               H-87
                                                                  OLM04.2

-------
Exhibit H -- Appendix A
Format of Records for Specific Uses
5.0   TYPE 90 -- COMMENTS RECORD

      This record could appear anywhere after the Record Type 10.   The comment
      on this record will usually apply to the most previous record type.   The
      example of Record Type 90 below would appear after  the  Record Type  10 -
      indicating the software and version number  that was used to generate  this
      data.
      90|CCS SYSTEMS SOFTWARE VERSION 3.5|000023A75
                                                                           \
                                     H-88                              OLM04.2

-------
                                       OLM04.2A
                     Corrections/Modifications/Clarifications
Corrections:
Exhibit C, Section 1.0 Volatiles Target Compound and contract Required Quantitation Limits

All values under the column "Med. Soil ug/Kg" should be changed from 1200 to 1300.  The reason for
this change lies in the fact that the CLP rounding rule has been changed from that used in previous
Statement of Work. The numbers in the "med. Soil" column are derived from calculations based on the
sample preparation protocol of the SOW'. The calculated "non-rounded" value under this column is
1250 ug/Kg.  Based upon the previous SOW rounding rule (if number to be retained is an even number
preceding a 5, then number is to be unchanged), 1250 ug/Kg would be rounded to 1200 ug/Kg. The
new rounding rule ( as outlined in Exhibit B, Section 3.3.9, page B-33) stipulates that if the number to
be dropped is 5 or greater, then the number retained should be increased by 1.  Based upon this  '
requirement, 1250 ug/Kg becomes 1300 ug/Kg.
Contract Modifications:

                              Exhibit D - Volatiles Appendix B
               Modified SW-846 Method 5035 for Volatiles in Low Level Soils
                                                                V
1.1    The following statement should be added to the end of the paragraph:  (It has been
Superfund's desire to make Method 5035 the predominant method for low level volatile soil analysis.
However, due to technical uncertainties and logistic problems, the method has so far had limited use.
AOC now believes that most of the technical and logistic issues have been resolved, at least on an
interim basis, and therefore will be requesting the Regions to specify the use of the CLP Method 5035.)

7.1.1  Delete 7.1.1 as written and insert the following:  Soil/sediment samples should be collected
in the field in either field core sampling/storage containers (i.e. Encore ™ or equivalent) or pre-
prepared closed-system purge-and-trap sample vials as described in Section 9.3. If field core sampling
containers are used, the field should send at least three containers per field sample containing
approximately 5 g each, and at least one 60 mL sealed glass vial containing sample with minimum
headspace.  The Contractor shall transfer the contents of the field core sampling containers immediately
upon receipt into the closed-system sample vial prepared as described in Section 9.3 below and record
the date and time of transfer. If pre-prepared closed-system purge-and-trap sample vials are used to
collect samples in the field, the field should send at least two closed-system containers per field sample
containing approximately 5 g each, and at least one 60 mL sealed glass vial containing sample with
minimum headspace. The Contractor is then to proceed to Section 9.3.9 and determine final sample
weight If me minimurn amount of containers have not been sent by the field, the Contractor is to
immediately contact SMO for instructions, note.  If MS/MSD analysis are required for a particular
sample, 2 additional field core sampling/storage containers or two additional closed-system purge-and-
trap sample vials should be sent by the field.  Contact SMO if insufficient sample for MS/MSD analysis
has been sent by the field.

-------
7.1.2   Change wording to include the following: All samples received in closed-system purge-and-
trap sample vials  must be iced or refrigerated at 4° C (+/- 2° Q from time of collection until time of
analysis.  Any sample received in field core sampling/storage containers that are not immediately
transferred to a closed-system purge-and-trap sample vial containing sodium bisulfate preservation
solution must be placed in a freezing compartment and held at
 -12 ° C (+/- 2 ° C) until time of analysis (see Section 9.3).

7.2.1    Add the following language: The samples must be protected from light and refrigerated at 4
0 C (+/- 2 ° C) ( -12  ° C , +/- 2 ° C, if placed in freezing compartment as stated in 7.1.2) from the time
of receipt until 60 days after delivery of a reconciled, complete sample data package to the Agency.
After 60 days, the samples may be disposed of in a manner that complies with all applicable
regulations.

9.3.1  Add the following note to the end of the paragraph: Note: There should be three field
core sampling/storage containers for each field sample. The contents of two of the field core containers
are to be processed using the steps outlined in 9.3.2 through 9.3.9. One of these prepared samples is
then to be used as the primary sample, while the other is to be used as a back-up sample, if necessary.
The contents of the third field core container shall be transferred to a tared dry closed-system purge-
and-trap  container (i.e. no preservative solution or stirring bar is to be added), weighed according to
9.3.9, and then stored in a freezing compartment at  -12 ° C (+/- 2 ° C).  This sample shall be used for
the medium concentration level methanol extraction procedure as described in Exhibit D-VOA Section
 10.1.4.2  if results of the original analysis indicate that medium level extraction is warranted.

9.3.8     Change the language for "Note" to the following: NOTE: Soil samples that contain
carbonate minerals may effervesce upon contact with the acidic preservative solution in the sample vial.
Therefore, if samples are known or suspected to contain high levels of carbonates, a test sample (from
the 60 mL glass vial) should be added to a clean vial and checked for effervescence. If a rapid or
vigorous reaction occurs, the Contractor may discard the test sample and proceed with sample
preparation by transferring the contents of each of the three field core sampling/storage containers into
separate  tared dry closed -system purge-and-trap vials (i.e. contains no preservative solution), two of
which should contain stirring bars.§Weigh the  vials as per 9.3.9, and then transfer all three vials to a
freezing  compartment at  -12 ° C (+/- 2 ° Q. If analysis is to begin immediately on the sample, the
contents of one of the field core sampling/storage containers may be transferred to a tared  closed-
system purge-and-trap vial containing 5 mL reagent water (no sodium bisulfate). The sample is then
analyzed by the procedure starting at 9.3.9.

9.4        Sample Purge-and-Trap

9.4.1      Change  the language to the following: Prior to sample purge, all soil/sediment samples
must be allowed to warm to ambient temperature. For those samples that have been stored in freezing
compartments and that will be analyzed by the  low concentration level protocol, 5 mL of reagent water
must be added to  the  vials without disturbing the hermetic seal of the sample vial (note: an additional 5
mL of reagent water  will be added to the vial as per 9.4.2).  Shake all vials containing aqueous
solutions gently, to ensure that the contents move freely and that stirring will be effective. Place the
sample vial in the instrument carousel according to the manufacturer's instruction.

-------
requested to use the preliminary results option judiciously. An attempt will be made to ensure that
appropriate turnaround options are requested and that unnecessary QC and reporting packages are not
performed. Under the new SOW, MS/MSD analysis is not required if the Agency requests that it not
be performed. The lab is requested to make sure that SMO is contacted if there is any question or
concern at all over assigning SDGs.

-------