PB83-247239
TEST PROCEDURE FOR  URANIUM  IN  DRINKING  WATER
INTERLABORATORY  COLLABORATIVE  STUDY
Monsanto Research  Corp.
Miamisburg, OH
Aug  83
                U.S. DEPARTMENT OF COMMERCE
             National Technical Information Service
                                            
-------
                                    P333-247239
                                         EPA-600/4-83-038
                                         August  1983
TEST PROCEDURE FOR URANIUM IN DRINKING WATER:
     Interlaboratory Collaborative Study
                     by
      C. A. Phillips and C.  T.  Bishop
Environmental  Assessment and Planning Section
   Mound Facility, Monsanto  Research Corp.
           Miamisburg,  Ohio   45342
     DOE Contract No.  DE-AC04-76-DP00053
               Project Officer

               Earl  Whittaker
         Quality Assurance Division
 Environmental Monitoring Systems Laboratory
          Las Vegas, Nevada  89114
       Contract No. EPA-IAG-79-D-X0736
 ENVIRONMENTAL MONITORING SYSTEMS LABORATORY
     OFFICE OF RESEARCH AND DEVELOPMENT
    U.S. ENVIRONMENTAL PROTECTION AGENCY
          LAS VEGAS, NEVADA  89114

-------
                                   TECHNICAL REPORT DATA
                            (Please read Instructions on the reverse before completing)
1. REPORT NO.

  EPA-600/4-83-038
                             2.
3. R£f I PINT'S ACCESSION NO
      '$3 -.-.-•£
                                  * '
4. TITLE AND SUBTITLE
 TEST  PROCEDURE FOR URANIUM IN DRINKING  WATER:
 Interlaboratory Collaborative Study
             5. REPORT DATE
              August  1983
             6. PERFORMING ORGANIZATION CODE
7. AUTHOR(S)
 C. A.  Phillips and C. T. Bishop
                                                           8. PERFORMING ORGANIZATION REPORT NO.
3. PERFORMING ORGANIZATION NAME AND ADDRESS
 Monsanto  Research Corporation
 Mound  Facility
 P.O. Box  32
 Miamisburg, Ohio  45342
                                                           10. PROGRAM ELEMENT NO.
             11. CONTRACT/GRANT NO.
                EPA-IAG-79-D-X0736
12. SPONSORING AGENCY NAME AND ADDRESS
 U.S.  Environmental Protection Agency—Las Vegas, NV
 Office  of  Research and Development
 Environmental  Monitoring Systems Laboratory
 Las  Vegas, Nevada  89114
             13. TYPE OF REPORT AND PERIOD COVERED
                Project. 4/80-9/80	
             14. SPONSORING AGENCY CODE
                EPA/600/07
15. SUPPLEMENTARY NOTES
 Mound  Facility is operated by Monsanto  Research Corporation for  the U.S.  Department of
 Energy under Contract No. DE-AC04-76-DP00053.
16. ABSTRACT
        An inter!aboratory collaborative study was conducted to test  a procedure selectee
 for  measurement of uranium in drinking water.  Drinking water samples containing 8.1,
 17.4,  and 75.3 pCi/1 were analyzed  in  triplicate by 18 participating laboratories.

      The statistical analyses of  the test results gave coefficients of variation for
 repeatability (within-laboratory  precision)  of 14.6, 8.1, and 8.3 percent for the three
 samples for an average repeatability precision of 10.3 percent  over the uranium
 concentration range of 8 to 75  pCi/1.   The analyses also gave coefficients of variatior
 for  reproducibility (combined within-  and between-laboratory precision) of 15.3, 14.9,
 and  9.1 percent for an average  reproducibility precision of 13.1 percent over the
 uranium concentration range of  8  to 75 pCi/1.

      The accuracy indexes of the  test  procedure for the three uranium concentrations
 was  98.0, 102.6, and 101.9 percent  for an average of 100.8 percent  over the uranium
 concentrations range of 8 to 75 pCi/1.
17.
                                KEY WORDS AND DOCUMENT ANALYSIS
                  DESCRIPTORS
                                              b.lDENTIFIERS/OPEN ENDED TERMS
                           c. COSATl Field/Group
18. DISTRIBUTION STATEMENT
 RELEASE  TO PUBLIC
                                              19. SECURITY CLASS (This Report)

                                                UNCLASSIFIED	
                           21. NO. OF PAGES
                               50
20. SECURITY CLASS (This page)

  UNCLASSIFIED
                           22. PRICE
EPA Form 2220-1 (R*v. 4-77)   PREVIOUS EDITION is OBSOLETE j

-------
                                   NOTICE

     This report has been reviewed in accordance with the U.S.  Environmental
Protection Agency's peer and administrative review policies and approved
for presentation and publication.  Mention of trade names or commercial
products does not constitute endorsement or recommendation for  use.

-------
                                   CONTENTS
Tables	    iv
Abbreviations and Symbols 	 ...     v
Acknowledgements	   vii

Introduction	     1
Conclusions 	     2
Recommendations 	     4
Experimental Procedures 	     6
     Analytical test procedure	     6
     Collaborative test procedures	     6
          Uranium standard solution preparation 	     6
          Uranium concentrate preparation 	     6
          General procedures	     8
     Data processing procedures 	     8
Results and Discussion	    13

References	    26

Appendices

     A.  Laboratories Participating in the Uranium in
           Drinking Water Collaborative Study 	    27
     B.  Analytical Test Procedure "Uranium in Drinking
           Water-Radiochemical Method, Method 908.0"	    29
     C.  Data Sheet	    36
     D.  Collaborative Study  Instructions 	    37
     E.  Questionnaire, Summary on Collaborative Study	    39
     F.  Comments from Participants in Collaborative Study	    40
                                    i i i

-------
                                    TABLES


Number                                                                    Page

  1       Uranium in Drinking Water, Replicate Analyses Data ......   17

  2       Uranium in Drinking Water, Precision Summary .........   20

  3       Uranium in Drinking Water, Cell Average (X-,-,-) and
            Cell Standard Deviation (S^-) ...... . .........   21

  4       Uranium in Drinking Water, Deviation of Lab Average from
            Grand Average (di-i, pCi/1) and Percent Deviation
            (X di) ..... •  .....................   22
  5       Uranium in Drinking Water, Reduced Deviations,
            (e1Jf pCi/D ........................   23

  6       Uranium in Drinking Water, Ratio of Lab Standard
            Deviation to Pooled Standard Deviation (k-j,-) ........   24

  7       Uranium in Water, Accuracy Index (A-,-) and
            Bias (tj) ............ .... ...........   25
                                     IV

-------
                       LIST OF ABBREVIATIONS AND SYMBOLS
ABBREVIATIONS

     ASTM
     CT
     c/m
     d/m
     DOE
     EPA
     g/cm3
     g/i
     IAG
     kg
     MCL
     mg/1
     nCi/g
     MIPDWR
     pCi/1
     ppm
     yCi/g
     2S %
The American Society for Testing and Materials
chloride
counts per minute
disintegrations per minute
The Department of Energy
Environmental  Protection Agency
grams per cubic centimeter
gram/liter
Interagency Agreement
kilogram
maximum contaminant level
milligram/liter
nanocurie/gram
National Interim Primary Drinking Hater Regulations
picocuries per liter
parts per million
microcurie/gram
Two sigma percent (two standard deviations in percent)
SYMBOLS
     %di

     n
     P
the accuracy index for the sample j level  of uranium
concentration
the deviation of Lab i average from the grand average
for sample j
the percent deviation of Lab i average from the grand
average for sample j
the reduced deviation of Lab i average from the grand
average for sample j
the repeatability interval for sample j level
the reproducibility interval for sample j level
the ratio of Lab i average standard deviation (Sij)
to the pooled standard deviation (Sr.) for sample j
the number of replicate analyses    '•'
the number of participants in the collaborative study
the standard deviation of the replicate test results
for sample j by Lab i
the standard deviation of between-laboratories
precision for sample j

-------
Sr.            — the repeatability (within-laboratory)  standard
  J               deviation for the sample j
$r.            — the smoothed repeatability standard deviation for
  ^               sample j
SR •            — the reproducibility (combined within-  and between-
  J               laboratory) standard deviation for sample j
!>R.            — the smoothed reproducibility standard  deviation for
  J               sample j
Sx.            — the standard deviation of the grand average  for
  J               sample j
tc             -- the critical value of the bias test for P
                  participants
tj             — the bias value for sample j
Tn             -- the T-test value of the nth replicate, outlier test
               — the adjusted within-laboratory standard deviation
               -- the coefficient of variation for between-laboratories
                  precision for sample j
Vr.%           ~ the coefficient of variation for repeatability
  J               (within-laboratory) for sample j
               -- the coefficient of variation for reproducibility
                  (combined within- and between-laboratory) for sample j
               — the test result of the replicate h
               -- the arithmetic average of all replicate test results
                  of sample j by Lab i
               ~ the test result of replicate h of sample j by Lab i
X~j°"           — the grand average value for sample j
Yi             -- the known value of the sample j uranium concentration
                  (pCi/1)
VLj%
VRj%

-------
                               ACKNOWLEDGEMENTS
     The authors would like to thank  all  of those  individuals  from the
participating laboratories involved in  the  collaborative  study (Appendix A).
Special thanks are due to Richard Velten  of the  U.S.  Environmental Protection
Agency's Environmental Monitoring and Support  Laboratory,  Cincinnati, Ohio,
who developed the radiochemical  method  used in this study.  Thanks are also
given to A. A. Glosby of Mound Facility for her  efforts  in  sample preparation
and packaging, to V. R. Casella and W.  H. Yanko  for many  helpful discussions,
and to K. N. McLennan for typing the  final  manuscript.
                                    vii

-------
                                 INTRODUCTION
     The National Interim Primary Drinking Water Regulations (NIPDWR)  imply
that when a gross alpha measurement of a drinking water sample exceeds 15
picocuries per liter (pCi/1),  then an analysis for uranium should  be done  to
determine the uranium alpha contribution to the gross alpha concentration.
The method listed in the NIPDWR for the measurement of uranium in  drinking
water is a fluorometric method which determines uranium in mass units.  It  is
now known (subsequent to the promulgation of the NIPDWR) that the  ratio of
uranium alpha activity to uranium mass concentration in ground water can vary
significantly from that which  is common to natural  uranium ore deposits.
Therefore, a test procedure for measuring total uranium alpha activity
concentrations is needed that  will more adequately relate to drinking  water
gross alpha activity.  Such a  test procedure was selected and tested in an
inter!aboratory collaborative  study.  This is a report of the results  of that
study.

     The method of analysis used in this study is a simplified version of  a
method that measures the uranium isotopic concentrations in the sample by
alpha spectrometry.  This method measures total uranium alpha activity, the
measurement needed for a gross alpha assessment of a drinking water sample.
This method is included in the updated EPA procedures manual, EPA-600/4-
80-032, August 1980, "Prescribed Procedures for Measurement of Radioactivity
in Drinking Water".

     The purpose of the collaborative study of the selected test procedure is
to determine what precision and accuracy can be expected by the use of the
procedure by any competent laboratory in the analysis of drinking  water
samples for alpha activity contributed by the uranium in the water samples.

-------
                                  CONCLUSIONS


     The repeatability precision (within-laboratory precision),  reproduci-
bility precision (the combined within- and between-laboratory precision),  and
the accuracy have been determined for the test procedure by a multilaboratory
test in this study, and they hereby become criteria by which to  evaluate an
alternate test procedure for equivalency.

     The estimated repeatability (within-laboratory),  single-operator,  single
machine, same-day, relative precision of the test procedure for  the
determination of total uranium alpha concentrations at the 8.1 pCi/1  level
(averaged over the 18 participants) is ±2.3 pCi/1 (28.4%, 2S%);  at the  17.4
pCi/1 level (averaged over the 17 participants) is ±2.9 pCi/1 (16.5%, 2S%);
and at the 75.3 pCi/1 level (averaged over the 18 participants)  is ±12.8 pCi/1
(17.0%, 2S%), for an average 2S percent of 20.6 percent for the  range of
uranium alpha concentrations of 8 to 75 pCi/1.

     The estimated reproducibility (combined within- and between-laboratory),
multi-operator (multilaboratory), single-machine, same-day, relative  precision
of the test procedure in the determination of total uranium alpha
concentrations at the 8.1 pCi/1 level  (averaged over the 18 participants)  is
±2.4 pCi/1 (29.6%, 2S%); at the 17.4 pCi/1 level  (averaged over  the 17
participants) is ±5.3 pCi/1 (30.4%, 2S%); and at  the 75.3 pCi/1  level
(averaged over the 18 participants) is ±14.0 pCi/1 (18.5%, 2S%); for  an
average 2S% of 26.2 percent for the range of uranium alpha concentrations  of
8 to 75 pCi/1.

     In terms of the 95 percent repeatability interval (on the basis  of test
error alone) the difference in absolute value, of two test results obtained in
the same laboratory at the 8.1 pCi/1 level of uranium alpha concentration  can
be expected to exceed 2.3 pCi/1 only 5 percent of the time; at the 17.4 pCi/1
level, 5.2 pCi/1 only 5 percent of the time; and  at the 75.3 pCi/1 level,  22.4
pCi/1 only 5 percent of the time.  These average  to be 29.3 percent of  the
uranium alpha concentration over the range of 8 to 75 pCi/1.  When differences
exceed 29.3 percent of the concentration, then one or both of the test  results
are suspect.

     In terms of the 95 percent reproducibility interval (on the basis  of  test
error alone), the difference, in absolute value,  of two test results  obtained
in different laboratories at the 8.1 pCi/1 level  of uranium alpha
concentration can be expected to exceed 2.9 pCi/1 only 5 percent of the time;
at the 17.4 pCi/1 level, 6.6 pCi/1 only 5 percent of the time; and at the  75.3
pCi/1 level, 28.5 pCi/1 only 5 percent of the time.  These average to be 37.2

-------
percent of the uranium alpha concentration  over the  range 8 to  75 pCi/1.  When
differences exceed the 37.2 percent  of the  uranium alpha concentration, then
one or both test results are suspect.

     The accuracy of the test procedure  expressed as the accuracy index in
percent for the three uranium alpha  concentration levels are 98.0, 102.6, and
101.9 percent respectively, for an average  accuracy  index of 100.8 percent
over the range of concentrations of  8 to 75 pCi/1.

     The sensitivity of the test procedure  is  not limited as much by the
chemistry of the procedure as it is  by the  background  and the counting
efficiency of the alpha counting system  used.   With  a  1-liter sample, an alpha
background of 0.1 cpm, a counting time of 100  minutes,  and a counting
efficiency of 15 percent (0.15 cpm/dpm), the sensitivity of the test procedure
for uranium alpha would be about 0.3 pCi/1  at  the 95 percent confidence level
(using sensitivity as defined in the NIPDWR).

     The lack of an adequate carrier with which to determine chemical recovery
in each sample analyzed, and having  to use  a predetermined recovery factor,
causes some uncertainty in the adjusted  test results (adjusted  by recovery
factor).  However, the predetermined recoveries averaged for the 18
laboratories was 91. ± 15. percent,  indicating a relatively high chemical
recovery for the test procedure and  therefore  minimal  uncertainty in the
adjusted test results.

     The t-test to show significant  differences and  method bias showed that
the test procedure did not contain significant systematic errors and therefore
no bias was indicated for uranium alpha  concentrations  up to 75 pCi/1.

-------
                               RECOMMENDATIONS


     It Is recommended that the test procedure used in this study (described
in detail in Appendix B) be used to monitor drinking water samples for uranium
alpha contribution to the gross alpha activity when the gross alpha activity
exceeds the amount specified in the NIPDWR (15 pCi/1, Section 141.15).

     It is recommended that a change be made in the preparation of the
separated uranium fraction for counting the alpha activity.  It is also
recommended that the change be verified experimentally to demonstrate the
benefit to the Test Procedure before the change is incorporated into the Test
Procedure.  To make that change in the preparation of the separated uranium
fraction for alpha counting, the following changes and additions to the Test
Procedure (Method 908.0) are proposed.  Replace existing steps 8.2.5 through
8.2.9 with the following 8.2.5 through 8.2.10' steps.

     8.2.5  Elute the uranium with six column volumes of 0.1N_HC1, collecting
            the eluate in a 150 ml beaker.

     8.2.6  Evaporate the eluate to near dryness, then add 1 ml 12N HC1
            (cone.), 10 ml of water, 0.2 ml 20% TiCIs, and 1 ml of lanthanum
            carrier solution, stir.

     8.2.7  Add 0.5 ml HF (cone.), stir well and allow to stand for 30
            minutes.

     8.2.8  Filter through 47 mm, 0.2 \im pore membrane filter, collecting the
            coprecipitated U/LaF3.

     8.2.9  Wash the U/LaF3 with 10 ml of water followed by 10 ml of
            ethanol.

     8.2.10 Air dry the filter for at least 1 hour before counting for alpha
            activity.

Also add the following item to the "Apparatus" section of the procedure.

     5.7    0.2 pm pore, 47 mm diameter membrane filter that withstand the
            acid treatment in the Test Procedure and will lay flat after
            drying (such as Gelman AN-200).

And add the following items to the "Reagents" section of the procedure.

     6.15   Lanthanum nitrate, (1.0 mg La+^/ml).  Disolve 3.11 g
            La(N03)3'6H20 in one liter of 0.1N_ HN03.

-------
     6.16   Hydrofluoric acid, 29N :  HF (cone.) sp. gr. 1.18, 49%.

     6.17   Titanium trichloride, TiCl3 :  20%

     The term "column volume" in the test  procedure should be clarified.   This
can be done by simply giving the milliliters of the resin bed volume specified
in the procedure.  A resin bed 1.3 cm in diameter by 8.0 cm high will have a
volume of 10.6 ml.  Then in Section 8.2.3  of the procedure, after the
expression "— with 6 column volumes", put in parentheses (6 x 10.6 ml =  63.6
ml, or 65. ml, rounded to the nearest 5 ml).

     Since no carrier or tracer is used in the procedure with each sample to
determine chemical recovery, it is recommended that with each set of samples
to be analyzed by this test procedure, a spiked sample (to determine recovery)
and a sample duplicate (to verify precision) be analyzed.  This recommendation
should be incorporated into the procedure.

-------
                            EXPERIMENTAL PROCEDURES
ANALYTICAL TEST PROCEDURE
     The analytical  test procedure used in this  study is  described  in  detail
in Appendix B.  The procedure is capable of quantifying all  uranium alpha
emitting isotopes in drinking water.


COLLABORATIVE TEST PROCEDURES

Uranium Standard Solution Preparation

     An ampoule of uranium standard solution,  prepared by the National  Bureau
of Standards, was sent to each participant for recovery and  counting
efficiency determinations.  That solution had  a  total  uranium (U-238,  U-235,
and U-234) alpha specific activity of 5,171 dpm/g.

Uranium Concentrate Preparation

     Three uranium concentrates were  prepared  for the collaborative study.
The concentrates were prepared by dilution of  a  standard  uranium solution  of
natural isotopic composition prepared by the National  Bureau of Standards
(NBS) for the Quality Assurance Division, U.S. Environmental  Protection
Agency, Las Vegas, Nevada.  The standard solution was one molar in  nitric
acid.  The calculated isotopic uranium radioactivity concentrations on
November 1978, as specified by the National Bureau  of Standards, were
uranium-234, 2,478 ± 129 disintegrations per minute per gram (dpm/g);
uranium-235, 119 ± 0.5 dpm/g, and uranium-238, 2,574 ± 8  dpm/g.

     Uranium concentrate #1 was prepared by delivering 0.6910 g of  the NBS
stock solution to a tared 1,000-milliliter volumetric flask  and diluting to 1
liter with 0.5 M^ HC1.  The final weight of the solution was  1,005.4 grams  and
the uranium concentration was calculated to be 3.55 ± 0.19 dpm/g.  The
specific gravity of the solution was  1.0067 g/ml  at 22.2°C based on the
average of two determinations using a calibrated pycnometer.  The concentra-
tion of the solution was then calculated to be 3.58 ± 0.19 dpm/ml.   A
5-milliliter sample of this solution  diluted to  1 liter with uranium-free
water is equivalent to a total uranium concentration of 8.1  ± 0.4 pCi/1.
After complete mixing of the solution, approximately 20 ml portions were
sealed in 25 ml glass ampoules labeled uranium concentrate #1 and sent to  the
participants for analysis.  Uranium concentrates #2 and #3 were prepared in
the same manner as uranium concentrate #1 and  data  relative  to all  three
concentrates are shown as follows:

-------
                                                                       Cone.
               Weight of    Total  Weight           Specific              When
  Uranium    NBS Standard   of 0.5 M HC1    Cone.    Gravity     Cone.    Diluted*
Concentrate  Standard (g)   Solution (g)   (dpm/g)   (g/ml)    (dpm/ml)   (pCi/1)
    #1


    n


    #3
0.6910


1.4924


6.4559
1005.4


1005.1


1005.3
                           3.55
                           ± 0.19

                           7.68
                           ± 0.40

                          33.2
                          ± 1.7
           1.0067
           1.0068
           1.0072
 3.58
 ± 0.19

 7.73
 ± 0.40

33.4
± 1.7
 8.1
 ± 0.4

17.4
± 0.9

75.3
± 3.9
*5 ml of the concentrate diluted to 1 liter.
     The concentrations of the three uranium concentrates  were  verified  by
alpha pulse-height analysis comparisons to a uranium-232 standard  which  had
been previously calibrated with NBS standardized uranium-236.   A 1-milliliter
aliquot of uranium-232 standard solution was added to  a  1-milliliter sample  of
each concentrate.  These solutions were mixed,  made 8  molar in  hydrochloric
acid, and passed through AG1-X4 anion exchange  resin to  remove  daughter
activities which were present.  The uranium was eluted with 0.1 M^ HC1  and
electrodeposited on a 0.75-inch diameter stainless steel  disk for  alpha
pulse-height analysis (APHA) spectrometry.  By  comparing the net count rate  of
the uranium-232 to the net count rates of uranium-234, -235, and -238, the
concentrations of the uranium concentrates were determined. The values  of the
concentrates fell within the error limits as derived from  the errors  specified
by the NBS for the standard solution.  The following data  summarizes  the
results of this verification.
  Uranium
Concentrate
    #3
Known Concentration
  Calculated Fran
  Dilution of NBS
     Standard
     (dpm/ml)	

   3.58 ± 0.19

   7.73 ± 0.40

  33.4  ± 1.7
  Concentration
Measured by APHA
    (dpm/ml)*

   3.57 ± 0.12

   7.62 ± 0.19

  34.4  ± 3.4
% Difference From
    Known Value

        0.3

        1.4

        3.0
*Errors are one sigma counting errors only.

                                   7

-------
General Procedures

     A total of 25 laboratories agreed to participate in the collaborative
study and 19 laboratories submitted their results.   All  laboratories  who
submitted results are listed in Appendix A.   All  laboratories were supplied
with the proposed EPA procedure for the analysis  of total  uranium alpha
activity in drinking water (Appendix B), a Data Sheet (Appendix C), a
Collaborative Study Instruction Sheet (Appendix D), a Questionnaire (Appendix
E), and three unknown reference samples labeled uranium concentrate 1, 2,  and
3.

     By diluting a 5-milliliter aliquot of each concentrate to 1 liter with
the participant's drinking water, the concentrations of the reference samples
were calculated to be 8.1, 17.4 and 75.3 pCi/1  total uranium alpha activity.

     Each participating laboratory was requested  to perform triplicate
analyses of a 1-liter sample of their drinking  water to determine a blank
value.  The authors believed that some of the participants might have had  a
high drinking water blank and therefore bias the  reference values.  A blank
value for this study was therefore defined as the counts per minute of 1 liter
of the participant's drinking water including counter background.

     The results of each laboratory were reported on a data sheet which
specified sample identification, volume, gross  counts, counting time, counter
background, total uranium alpha activity in pCi/1,  average recovery factor
(R), and average counting efficiency (E).  All  calculated data were checked at
Mound Facility, and any significant differences between the calculated and
reported values were resolved with the individual participants.


DATA PROCESSING PROCEDURES

     The data from the 19 participants were tested for outliers by the ASTM
recommended criterion for rejection (ASTM 1980).   For this rejection
criterion, with n observations listed in order  of increasing value by xj _<
X2 1 X3 .1 ••• .1 xn» ^ tne outlying values xn or  xj are in question,
then Tn or Tj were calculated as follows:


                              Tn  =  (xn - x)/s

                                         - x)/s
     where:     Tn; Tj  =  test criterion

                     x  =  arithmetic average of all  n values

                     s  =  the estimate of the population standard deviation
                           based on the sample data

-------
If Tn or TI values exceeded the critical  value,  the measurement  in
question was rejected.  Critical  values of T for a 5 percent  two-sided  level
of significance were used for the rejection criterion.

     A statistical evaluation of the test results was carried out by  the
procedures described in E-691, E-177, and E-178  of the  ASTM Standard  Part  41,
1980.  The standard deviations and other statistical  parameters, and  equations
for their calculations are listed below.   The standard  deviation of individual
participant (or laboratory) test results, S-jj, was determined by Equation  2.
                     sij
                        1/2
                                            (2)
h=l
where:  X-jj^  =  the result reported for the h replicate of  the j
                 sample material  by Lab i

         R-jj  =  the mean of the individual  results  for sample  j  for Lab  i

         nn-j  =  the number of replicates reported for sample j by  Lab  i


The repeatability (within-laboratory) standard deviation, Sr.,  for  each
sample was determined by Equation 3 and 3A.                  J

     Where the number of replicates was the  same (3)  for all participants,
Equation 3 was used.
                                                                          (3)
      where:  P  =  the number of participants in the study.

     Where the number of replicates were not  the same for all  participants,
Equation 3A was used.
                                                                         (3A)

-------
The standard deviation of the grand average for each sample,  S%.,
was determined by Equation 4.                                  J
   p

  i=]
                                                  1/2
                                                                          (4)
      where:  R-JJ  =  the average of the test results  for sample
                      material  j by Lab i

               Xj  =  the grand average for sample material  j

The standard deviation of between-laboratories precision for each  sample
material, SL., was determined by Equation 5.
            J
                2
               -  /n
                                              \l/2
                                                                          (5)
The reproducibility (combined within- and between-laboratory)  standard
deviation for each sample, SR., was determined  by  Equation  6.
S
R   =  (Srj2 + SLj2)l/2
                                                                          (6)
The deviation of laboratory averages  from the  grand  average  for  each  sample,
dij, was determined by Equation 7.





The percent deviation of laboratory averages from the  grand  average for each
sample, % dij, was determined by Equation 8.
      =  100
                                                                          (8)
The estimated standard error of a cell  average,  calculated on the  basis  of
replication error only, U j ,  for sample  #1,  is  determined  by  Equation  9.
                             =  S
                                               (9)
                                    10

-------
For samples #2 and #3, duplicate analyses were made by one lab.   Thus  for
samples #2 and #3 the nP value (total  replicates)  in Equation 9  was  replaced
by  P
Reduced deviations (e-jj) for each participant for each sample were
determined by Equation 10.
                                                                         (10)
The ratio (kij) of the standard deviation of participant  replicate  analyses
(Sjj) to the pooled standard deviation (Sr.) for each sample was
determined by Equation 11.                J


                               kij  =  Sij/Srj                           (11)


The coefficient of variation for repeatability (within-laboratory)  for each
sample, Vr.%, was determined by Equation 12.


                              Vrj%  =  100 Srj./Xj                        (12)


The coefficient of variation for between-laboratories precision for each
sample, Vi-%, was determined by Equation 13.
          J

                             vLj%  =  100 SLJ/XJ                         (13)


The coefficient of variation for reproducibility (combined within-  and between
laboratory) for each sample, VR.%, was determined by Equation 14.
                               J

                             VRj%  =  100 SRj/Xj                         (14)


Smoothed values of the repeatability and reproducibility  standard deviations
for each sample, S"r. and SR. were determined by Equations 15 and 16,
respectively.      J       J


                           §ri  =  (vV%)(Kj)/100                         (15)
                             J
                           SR.J  =  (Wj/ioo                           (16)
                                     11

-------
     Where 7r%  =  average of Vr.% for the 3 samples
                                J
           V~R%  =  average of VR.% for the 3 samples
                                J
Assuming that the test errors are normally distributed and using a 95 percent
probability level, the 95 percent repeatability interval  (Irj) and the 95
percent reproducibility interval (iRj) for each sample were determined using
the smoothed standard deviations in Equations 17 and 18,  respectively.


                IP-  =  2(2)1/2 Sr.  =  2.83 Sr.                         (17)
                  J               J            J


                IR.  =  2(2)1/2 L.  =  2.83 Sr.
                  J               J            J

The accuracy index (Aj%) a percent relationship of the grand average
(Xj) to the known value (Yj) for each sample material  was determined
by Equation 19.


                            Aj%  =  100 Xj/Yj

A comparison of the grand average value (Xj) with the known value
(Yj) for each sample in a bias test was determined by Equation 20 (Youden
and Steiner, 1975).


                          ?. Y-
                           J  J                                          f?ni
                 tj  =  	   ,,y  , (P-l) degrees of freedom           {  u>

                         xj

   where tj  =  the bias test result for the j sample

          P  =  the number of participants
                                    12

-------
                            RESULTS AND DISCUSSION


     For this collaborative study of the selected test procedure (Appendix B),
uranium in water concentrate solution batches were prepared and a portion was
sent to each participant.  Three separate batches were prepared with three
different uranium concentrations.  Concentrates were used so that participants
would make specified dilutions with their own drinking water, thereby
subjecting the test procedure to 19 different drinking water sources.

     The test results from the 19 participants for the three uranium
concentrations in drinking water are tabulated in Table 1.  The participants
are identified by randomly assigned laboratory numbers.  Table 1 also lists
the counting error (1 sigma, Su) associated with each test result.  The
recovery data in Table 1 is a listing of the three (and sometimes four)
replicate uranium analyses of spiked (with standard uranium) samples by the
test procedure by each laboratory for the purpose of recovery determination.
The recovery factor (average of the replicate tests) was used to correct
analytical results of the three test samples.

     The test results of Lab 12 for Sample 2 and Lab 20 for samples 1 and 3
failed the laboratory outlier test at the 5 percent significance level  (ASTM,
E-178, 1980) and were therefore rejected (not used in the precision and
accuracy calculations).  The test results of Lab 20 for Sample 2 did not fail
the outlier test but was found to be significantly different from the grand
average by the t-test for significant differences (Youden and Steiner,  1975)
and was therefore rejected (not used in the precision and accuracy
calculations).

     Table 2 gives the known value (Yj) and grand average (Rj) for
each sample and the following precision parameters.

Sj».,  the standard deviation of the grand average for the j sample
  J
Sr.,  the pooled standard deviation (within-laboratory) for the j sample
  J
St.,  the square root of the component of variance between-!aboratories,
      CO
       Lj 5 for the j sample

SR.,  an estimate of the between-1aboratories precision, a combined
  J   within- and between-laboratory precision, or reproducibility precision
      for the j sample
                                     13

-------
Vr.%, the coefficient of variation for repeatability,  or  the  estimated
  J   relative standard deviation for within-laboratory precision,  or
      repeatability index for the j sample

V|_.%, the coefficient of variation for between-laboratories precision for
  J   the j sample

Vp.%, the coefficient of variation for reproducibility, the percent
  J   coefficient of variation for the total  variability  of a single test
      result including both within- and between-laboratory variability,  or
      reproducibility index for the j sample

S"r.,  the smoothed value of Sr.

So.,  the smoothed value of SR.
  J                           J
Ir.,  the repeatability interval—the maximum permissible difference due to
  J   test error between two test results  obtained  in  a single laboratory (or
      by a single participant) at the j sample level

ID.,  the reproducibility interval—the maximum permissible difference due
  J   to test error and systematic error between two test results obtained in
      two different laboratories at the j  sample level

     Table 3 lists, by participant, the average and the standard deviation of
the participant test results for each of the  three  sample materials (cells).
Table 3 also lists the Sr. and an adjusted Sr. value designated by  Uj for
for each sample material.J                   J

     Table 4 lists, by participant, the deviation and  the percent deviation of
the participant average from the grand average for  each sample material  and
the standard deviation of the grand average for each sample material,

S*j'
     Table 5 lists, by participant, the reduced deviations (e-j) for each
sample material.

     Table 6 lists, by participant, the ratio (k-j)  of  the participant test
results standard deviation (Sjj) to the pooled standard deviation (Sr-)
for each sample material.  This table points  out potential outliers. JSample 1
for Participant 8, Sample 2 for Participant 2, and  Sample 3 for Participant
14, show ratio values higher than the critical value for  17 or 18 participants
and 3 replicates (2.06).  Since the average of those participant and sample
test results were not significantly different from  the respective grand
average, those test results were not rejected.

     For each of the three sample materials,  Table  7 lists the known value
(Yj, pCi/1); the grand average, (Kj, pCi/1);  the accuracy index,  (Aj, %);
the standard deviation of the grand average,  (S^,  pCi/1); the bias
value, (tj); the critical value for significance for 17 and 18 participants,
(tc); and the number of participants in the study,  (Pj).  The table also
                                     14

-------
shows the average accuracy index for the test  procedure  (100.8%).   Table  7
shows no bias in the test procedure for the range of uranium concentrations
tested (8 to 75 pCi/1).

     The average of the recoveries for the 18  participants  is 91 ±  15  percent,
indicating that the chemistry of the test procedure is quite accurate  and
reliable.

     From Table 2 it can be seen that the standard deviations (Srj, SM,
and SRJ) depend on the level  of uranium concentration.   However, the percent
coefficients of variation as calculated from those standard deviations (Vr.
%, VM %, and VRJ %) show a rather fixed percent relationship to all threej
levels of uranium concentration.

     The repeatability (within-laboratory) and reproducibility (total
between-laboratory) intervals (Irjand IRJ) for each sample material were
calculated with smoothed Srj and SRJ values (Sri and SRJ,  respectively).  The
values listed in Table 2 were calculated at the 95 percent  probability level
and are thereby designated as the 95 percent repeatability  and 95  percent
reproducibility intervals.  The 95 percent repeatability interval  is that
difference value based on test error alone between two test results obtained
by the same participant (or laboratory) that would be exceeded only about 5
percent of the time.  The 95 percent reproducibility interval is that
difference value between two test results obtained by two  different partici-
pants (or laboratories) that would be exceeded only about  5 percent of the
time.  When difference values exceed the interval values,  one or both  test
results are suspect.  The study shows the estimated average 95 percent
repeatability interval to be 29.3 percent of the uranium concentration level
over the range of 8 to 75 pCi/1.  The estimated average  95  percent  reproduci-
bility interval is shown to be 37.2 percent over the same  uranium  concentra-
tion range.  It should be realized that those interval  values are  based on a
limited 18 participant average.

     In the questionnaire sent to all of the participants  in the study,
Question 5 asked, "Do you believe that this would be a  good reference  method
for uranium in drinking water?"  Fifteen participants answered "yes",  two said
"no", and two were undecided.  Appendix E is a summary  of  responses to the
questionnaire.

     Appendix F is a listing of the comments made by the participants  about
the test procedure.  Responses to Questions 2, 5, 6, and 7  are addressed  in
the Recommendations section of this report.  In response to Question 3 one
participant commented that the method of transfer of the uranium eluant from
the beaker to the counting planchet was questionable.  Such transfers  are made
in many radiochemical methods and should not cause serious  problems.   In
response to Question 4 one participant suggested that new  ion exchange resin
should be specified for each analysis rather than regenerating the  resin.
This can and should be done if the analyst does not have confidence in the
regenerating procedure.  The regenerating procedure should  completely
regenerate the resin from drinking water levels of uranium  concentrations.
Re-use of the resin is not recommended if high levels of uranium concentration
(>500 pCi/1) have been put through the resin bed.  In response to  Question 10


                                     15

-------
one participant commented that the uranium yields would vary with  the
dissolved solids in the water sample.   High carbonate content is usually
associated with high dissolved solids  and carbonate does complex uranium and
tends to prevent its hydroxide precipitation.   The test procedure  addresses
this problem adequately.  The 19 different water supplies used in  this
collaborative study suggest that uranium yield is not tied to the  dissolved
solids content when this test procedure is followed.
                                     16

-------
        TABLE 1.   URANIUM IN DRINKING WATER,  REPLICATE  ANALYSES  DATA
Lab
1
2
3
4
5
7
8
9
10
Sample Uranium
1
8.67 CO. 335)
9.42 (0.349)
9.18 (0.345)
10.46 (0.641)
7.85 (0.558)
7.69 (0.553)
7.02 (0.432)
7.69 (0.450)
7.97 (0.458)
9.12 (0.809)
8.23 (0.771)
8.02 (0.761)
8.17 (0.298)
7.89 (0.293)
7.62 (0.288)
6.13 (0.367)
6.40 (0.373)
7.11 (0.388)
3.97 (0.888)
11.01 (1.341)
8.28 (1.186)
8.31 (0.742)
7.05 (0.689)
7.17 (0.694)
9.22 (0.602)
8.87 (0.591)
8.95 (0.594)
Concentration, (counting
2
22.76 (0.537)
22.51 (0.534)
19.57 (0.499)
25.01 (0.981)
19.94 (0.877)
18.72 (0.851)
13.66 (0.587)
13.33 (0.580)
12.05 (0.553)
. 14.62 (1.016)
15.44 (1.043)
(15.0)
17.61 (0.432)
15.59 (0.407)
17.18 (0.426)
17.71 (0.681)
18.52 (0.690)
18.86 (0.693)
21.05 (1.800)
21.92 (1.834)
22.92 (1.873)
17.22 (1.043)
18.73 (1.086)
16.68 (1.027)
17.88 (0.833)
17.49 (0.824)
17.53 (0.825)
error), a pCi/1
3
93.03 (1.081)
84.07 (1.027)
75.21 (0.972)
92.06 (1.872)
74.13 (1.680)
86.83 (1.818)
83.88 (1.418)
75.37 (1.345)
76.86 (1.358)
73.38 (2.252)
76.47 (2.298)
(74.9)
74.39 (1.242)
79.29 (1.282)
75.48 (1.251)
74.84 (1.133)
82.85 (1.183)
73.72 (1.126)
64.98 (3.090)
75.32 (3.321)
80.48 (3.431)
76.60 (2.160)
69.25 (2.055)
71.24 (2.083)
70.16 (1.642)
72.65 (1.671)
66.52 (1.599)
Recovery &
Percent
81.43
76.23
91.08
78.2
69.1
80.0
101.0
97.3
106.0
82.5
79.5
102.5
97.6
96.1
83.0
83.0
89.9
45.0
36.0
41.0
115.5
104.9
108.3
99.0
96.0
99.0
                                                                 (continued)
See footnotes at end of table, page 18.

                                     17

-------
                             TABLE 1.   (Continued)
Sample Uranium Concentration, (counting error), a pCi/1
Lab
11
12
14
18
20
21
22
23
1
8.16 (0.525)
6.76 (0.482)
7.41 (0.503)
8.38 (0.239)
7.70 (0.230)
6.78 (0.217)
7.54 (0.721)
8.46 (0.755)
8.84 (0.769)
6.97 (0.662)
9.11 (1.141)
7.12 (0.668)
5.29 (0.348) e
4.33 (0.316) e
5.47 (0.354) e
7.10 (0.367)
6.18 (0.344)
6.65 (0.356)
8.08 (0.436)
8.44 (0.446) .
7.44 (0.270)
7.98 (0.309)
7.98 (0.309)
8.01 (0.309)
2
14.67 (0.691)
17.05 (0.743)
15.72 (0.715)
2.79 (0.148) d
3.23 (0.157) d
3.14 (0.156) d
17.57 (1.032)
16.21 (0.995)
19.41 (1.079)
18.83 (1.018)
17.82 (1.550)
20.09 (1.640)
11.96 (0.518) e
12.20 (0.524) e
10.33 (0.483) e
15.95 (0.536)
17.25 (0.874)
17.60 (0.883)
16.73 (0.620)
18.10 (0.645)
17.02 (0.626)
16.24 (0.438)
17.30 (0.451)
17.24 (0.451)
Recovery b
3 Percent
83.22 (1.614)
88.80 (1.666)
65.67 (1.435)
76.84 (0.696)
77.85 (0.701)
77.35 (0.699)
81.61 (2.130)
92.51 (2.265)
64.37 (1.898)
73.28 (1.946)
80.15 (3.206)
78.64 (3.176)
56.64 (1.122) d
68.99 (1.731)
77.57 (1.835)
70.54 (1.750)
77.83 (1.073)
85.21 (1.120)
76.72 (1.065)
69.18 (0.898)
70.74 (0.908)
71.50 (0.791)
96.9
92.6
102.9
91.6
90.3
89.9
77.6
82.9
109.0
129.0
95.3
81.3
97.2
109.6
106.0
99.9
91.8
87.9
105.3
90.1
76.8 c
93.1
92.6
96.9
95.6
                                                                (continued)
See footnotes at end of table, page 18.
                                     1R

-------
                             TABLE 1.   (Continued)
          Sample Uranium Concentration,  (counting error),  a  pCi/1
Lab
24
25
1
8.18 (0.602)
6.23 (0.528)
7.12 (0.563)
9.17 (0.623)
7.47 (0.567)
10.83 (0.672)
2
21.72 (0.921)
21.99 (0.927)
17.88 (0.837)
18.42 (0.863)
16.52 (0.820)
14.70 (0.776)
3
70.50 (1.761)
69.43 (1.748)
71.14 (1.769)
81.44 (1.784)
83.91 (1.811)
85.31 (1.826)
Recovery D
Percent
101.7
91.6
104.3
93.7
' 93.0
111.1
a Su, counting uncertainty, pCi/1,  one sigma.

b R, chemical recovery using a spiked (standard)  sample,  the  average  of the
  three replicates for each lab was used as  factor (R)  in calculating the
  uranium concentrations in samples 1,2, and  3.   The  recoveries  given here do
  not correspond to the sample data listed,  but were determined independently
  with spiked samples by the test procedure.

c This chemical  recovery factor was used only  in  calculating  the  uranium
  concentration for sample 3.  The  average of  the other three values  was used
  to calculate the uranium concentration of  samples 1  and 2.         '

d Rejected by the ASTM Outlier Test.

e Rejected by t-test for significant differences  (Youden  and  Steiner, 1975).
                                     19

-------
            TABLE 2.  URANIUM IN DRINKING WATER,  PRECISION  SUMMARY
Parameter a
Yj (pCi/1)
Xj (pCi/1)
SXj (PCi/D
Srj (pCi/1)
SLj (PCi/D
SRj (pCi/1)
vrjx
V
VRj%
5rj (pCi/1)
sRj (pci/i)
Irj (pCi/l)
iRj (PCI/D
p

1
8.1 ± .4
7.9
0.76
1.15
0.37
1.21
14.6
4.7
15.3
0.82
1.04
2.31
2.93
18
Sample
2
17.4 ± .9
17.9
2.39
1.44
2.24
2.66
8.1
12.5
14.9
1.84
2.34
5.22
6.63
17

3
75.3 ± 3.9
76.8
4.6
6.39
2.83
6.98
8.3
3.7
9.1
7.92
10.06
22.43
28.47
18
Average






10.3
7.0
13.1





a Terms defined in text and the List of Abbreviations  and  Symbols.
                                     20

-------
TABLE 3.  URANIUM IN DRINKING WATER, CELL AVERAGE (Xjj)
           AND CELL STANDARD DEVIATION (S)
Sample
1

2

3

*„ S,j *„ Sll *,-J S,j
1 9.1
2 8.7
3 7.6
4 8.4
5 7.9
7 6.5
8 7.7
9 7.5
10 9.0
11 7.4
12 7.6
14 8.3
18 7.7
21 6.6
22 8.0
23 8.0
24 7.2
25 9.2
Xj 7.9
Srj 1.15
Uj 0.65
0.38
1.56
0.49
0.58
0.28
0.51
3.55
0.70
0.18
0.70
0.80
0.67
1.19
0.46
0.51
0.02
0.98
1.68



21.6
21.2
13.0
15.0
16.8
18.4
22.0
17.5
17.6
15.8
0
17.7
18.9
16.9
17.3
16.9
20.5
16.5
17.9
1.44
0.82
1.77
3.34
0.85
0.58
1.06
0.59
0.94
1.06
0.21
1.19
0
1.60
1.14
0.87
0.72
0.60
2.30
1.86



84.1
84.3
78.7
74.9
76.4
77.1
73.6
72.4
69.8
79.2
77.3
79.5
77.4
72.4
79.9
70.5
70.4
83.6
76.8
6.39
3.62
8.91
9.22
4.54
2.18
2.57
4.98
7.89
3.80
3.08
12.07
0.50
14.19
3.61
4.57
4.61
1.18
0.86
1.96



     0.76                2.39              4.65
                          9.1

-------
TABLE 4.  URANIUM IN DRINKING WATER, DEVIATION OF LAB AVERAGE FROM
   GRAND AVERAGE (djj, pCi/1) AND PERCENT DEVIATION (% d^)
Sample
Lab
1
2
3
4
5
7
8
9
10
11
12
14
18
21
22
23
24
25

dij
1.2
0.8
-0.3
0.5
0.0
-1.4
-0.2
-0.4
1.1
-0.5
-0.3
0.4
-0.2
-1.3
0.1
0.1
0.7
1.3
1
* dij
15.2
10.1
3.8
6.3
0.0
17.7
2.5
5.1
13.9
6.3
3.8
5.1
2.5
16.4
1.3
1.3
8.9
16.4

-11
3.8
3.4
-4.8
-2.8
-1.0
0.6
4.2
-0.3
-0.2
-2.0
0
-0.1
1.1
-0.9
-0.5
-0.9
2.7
-1.3
2
% dij
21.3
19.1
27.0
15.7
5.6
3.4
23.6
1.7
1.1
11.2
0
0.6
6.2
5.0
2.8
5.0
15.2
7.3

dij
7.3
7.5
1.9
-1.8
-0.4
0.3
-3.2
-4.4
-7.0
2.4
0.5
2.7
0.6
-4.4
3.1
-6.3
-6.4
6.8
3
%dij
9.5
9.8
2.5
2.4
0.5
0.4
4.2
5.7
9.1
3.1
0.6
3.5
0.8
5.7
4.0
8.2
8.3
8.8
         0.58
5.72
21.59
         0.76
2.39
 4.65
                               22

-------
                 TABLE 5.   URANIUM IN DRINKING WATER,  REDUCED
                        DEVIATIONS, (e-jj  a,  pCi/1)
Sample
Lab 1
1 1.9
2 1.2
3 -0.5
4 0.9
5 0.0
7 -2.1
8 -0.2
9 -0.6
10 1.7
11 -0.7
12 -0.4
14 0.6 '
18 -0.2
21 -1.9
22 0.1
23 0.1
24 1.1
25 1.9
Uj 0.65
Sr. 1.15
2
4.6
4.1
-6.0
-3.5
-1.3
0.6
5.0
-0.4
-0.3
-2.5
0
-0.2
1.3
-1.2
-0.7
-1.2
3.3
-1.6
0.82
1.44
3
2.0
2.1
0.5
-0.5
-0.1
0.1
-0.9
-1.2
-1.9
0.7
0.2
0.8
0.2
-1.2
0.9
-1.7
-1.8
1.9
3.62
6.38
a Term defined in text and the List of Abbreviations and Symbols.

-------
          TABLE 6.   URANIUM IN DRINKING  WATER,  RATIO  OF  LAB  STANDARD
                DEVIATION  TO POOLED  STANDARD  DEVIATION,  kj   a
Lab
1
2
3
4
5
7
8
9
10
11
12
14
18
21
22
23
24
25

1
0.33
1.35
0.42
0.51
0.24
0.44
3.08
0.60
0.16
0.61
0.70
0.58
1.04
0.40
0.44
0.01
0.85
1.46
Sample
2
1.23
2.31
0.59
0.40
0.74
0.41
0.65
0.74
0.15
0.83
-
1.11
0.79
0.60
0.50
0.41
1.59
1.29

3
1.40
1.44
0.71
0.34
0.40
0.78
1.24
0.60
0.48
1.89
0.08
2.22
0.56
0.72
0.72
0.18
0.14
0.31
a Term defined in text and the List  of Abbreviations and Symbols.

kc = 2.06 for 18 labs and 3 replicates

-------
         TABLE 7.  URANIUM IN WATER,  ACCURACY INDEX (Aj)  AND BIAS (tj)


                    	Sample	
Parameter a            1                 2                3            Average


 Yj (pCi/1)        8.1 ±  .4        17.4 ± .9        75.3  ±  3.9

 Xj (pCi/1)        7.9             17.9             76.8

 Aj %             98.0            102.6            101.9               100.8

 S*- (pCi/1)       0.76             2.39             4.65

 tj               -0.90             0.79             1.30

 tc                2.10             2.11             2.10

 P                18               17               18



a Terms defined in text and the List  of Abbreviations and Symbols.

-------
                                  REFERENCES
American Society for Testing and  Materials,  1980 Annual Book of Standards,
     Part 41, Designation:   E177-71  (Reapproved 1980) p. 195; E178-80, p. 206;
     and E691, p. 959.

Youden, W. J. and E. A.  Steiner,  "Statistical Manual of the Association of
     Official Analytical  Chemists",  Association of Official Analytical
     Chemists, Washington,  DC,  1975.

Krieger, H. L., "Interim Radiochemical Methodology for Drinking Water",
     EPA-600/4-75-008 (Revised) U.S.  EPA  Cincinnati, Ohio, 1976.

Krieger, H. L. and E. L.  Whittaker,  "Prescribed Procedures for Measurement of
     Radioactivity in Drinking  Water", EPA-600/4-80-032 August 1980, U.S. EPA
     Cincinnati, Ohio,  1980.

"National Interim Primary Drinking Water  Regulations".  U.S. EPA, Office of
     Water Supply, 401  M Street S.W., Washington DC, EPA-570/9-76-003.

Bishop, C. T., V. R. Case!la and  A.  A. Glosby,  "Radiometric Method for
     Determination of Uranium Water:  Single-Laboratory Evaluation and
     Interlaboratory Collaborative Study", EPA-600/7-79-093, U.S.
     Environmental Protection Agency, Las Vegas, NV, 1979.

-------
                                  APPENDIX  A
                       LABORATORIES  PARTICIPATING  IN  THE
                 URANIUM IN DRINKING WATER COLLABORATIVE  STUDY
Alabama State Department of Public Health
Montgomery, Alabama

Arkansas Department of Health
Little Rock, Arkansas

California Department of Health Services
Sanitation and Radiation Laboratory
Berkeley, California

Florida Department of Health and Rehabilitative Services
Radiological Health Services
Orlando, Florida

Hazen Research, Inc.
Golden, Colorado

Massachusetts Department of Environmental  Quality  Engineering
Lawrence Experiment Station
Lawrence, Massachusetts

Michigan Department of Public Health
Division of Radiological Health
Lansing, Michigan

New Hampshire Water Supply and Pollution Control Commission
Concord, New Hampshire

Oregon Department of Human Resources
Health Division Radiation Section
Portland, Oregon

Orlando Laboratories, Inc.
Orlando, Florida

Rockwell Hanford Operations
Richland, Washington
                                     27

-------
U. S. Army Environmental  Hygiene Agency
Aberdeen Proving Ground,  Maryland

U. S. Environmental Protection Agency
Environmental Monitoring  and Support Laboratory
Cincinnati, Ohio

U. S. Environmental Protection Agency
Eastern Environmental  Radiation Facility
Montgomery, Alabama

U. S. Food and Drug Administration
Winchester Engineering and Analytical  Center
Winchester, Massachusetts

University of Pittsburgh
Graduate School of Public Health
Pittsburgh, Pennsylvania

Union Carbide Corporation
Nuclear Division
Oak Ridge, Tennessee

Union Carbide Corporation
Nuclear Division, Analytical  Dept.
Paducah, Kentucky

Washington State Department of Social  and  Health  Services
Seattle,.Washington
                                     28

-------
                                  APPENDIX  B

                URANIUM IN DRINKING WATER-RADIOCHEMICAL  METHOD
                                 METHOD 908.0
                                  MAY 8, 1980

1.  Scope and Application

    1.1   This method covers the measurement  of total  uranium alpha  particle
          activity in drinking water.  Most drinking water sources,  especially
          ground water sources, contain soluble carbonates and  bicarbonates
          which complex and keep uranium in the water in solution.

    1.2   Uranium isotopic abundances in drinking water  sources are  apt  to  be
          present in ratios different from  the ratios in the deposits  from
          which the uranium entered the water sources.  The two predominant
          natural alpha emitting isotopes of  uranium are uranium-234 and •
          uranium-238.  Uranium-238 is the  predominant mass abundant isotope;
          greater than 99% compared to about  0.006% for  uranium-234.  However,
          uranium-234 has a specific activity for alpha  particle emission that
          is 1.8 x 101* times greater than that of uranium-238.   For  an equili-
          brium condition, the activity of  the uranium-234 is equal  to that  of
          the uranium-238.  Therefore, the  uranium mass  concentration  in water
          is not related to the alpha particle activity  of the  water.

    1.3   The Drinking Water Regulations under the Safe  Drinking Water Act,  PL
          93-523, require a measurement of  uranium for drinking water  samples
          that have a gross alpha activity  greater than  15 pCi/1. A mass
          uranium concentration measurement of a water sample cannot be
          converted to uranium alpha activity without first analyzing  for
          isotopic abundances.  Therefore,  a  method such as this one is  needed
          to determine the total uranium alpha activity  of the  sample, without
          doing an isotopic uranium analysis.

2.  Summary of Method

    2.1   The water sample is acidified by  adding HC1  and the sample is  boiled
          to eliminate carbonate and bicarbonate ions.  Uranium is coprecipi-
          tated with ferric hydroxide and separated from the sample.  The
          uranium is then separated from other radionuclides which were
          carried down with the ferric hydroxide by dissolving  the hydroxide
          precipitate in 8N^ HC1; putting the solution through an anion
          exchange column; washing the column with 8j^ HC1; and  finally,
          eluting the uranium with 0.1N HC1.   The uranium eluate is  evaporated
          and the uranium chemical form is  converted to  nitrate. The  residue
          is transferred to a stainless steel  planchet,  dried,  flamed, and
          counted for alpha particle activity.

-------
    2.2   Uranium recovery is determined with blank samples spiked with known
          amounts of uranium and taken through the procedure as  a regular
          sample.

    2.3   Counting efficiency is determined by transferring measured aliquots
          of a uranium standard to a planchet, diluting with 6-8 ml  of a 1
          mg/ml  HI03 solution in 4N^HN03, evaporating to dryness, flaming the
          planchet, and counting in an alpha counter.

3.  Sample Handling and Preservation

    3.1   Although carbonate ions in a water sample will  help to keep uranium
          in solution, the addition of extra carbonate or bicarbonate ions  to
          the sample as a preservative is not recommended because an increased
          carbonate concentration in the sample may cause some precipitation.
          Therefore, it is recommended that samples be preserved with HC1 to
          pH 2 at the time of collection.

    3.2   A sample size of, at least, 1 liter should be collected for uranium
          analysis.

4.  Interferences

    4.1   The only alpha-emitting radionuclide that may come through the
          chemistry and cause interference would be protactinium-231.
          However, protactinium-231 results from the decay of uranium-235,  a
          low abundance natural  isotope of uranium, and would therefore cause
          only a very small interference.

    4.2   Since uranium is a naturally occuring radionuclide, reagents must be
          checked for uranium contamination by analyzing a complete reagent
          blank by the same procedure as used for the samples.

5.  Apparatus

    5.1   Gas-flow proportional  counting system or

    5.2   Scintillation detection system.

    5.3   Electric hot plate.

    5.4   Ion exchange column:  approximately 13 mm (i.d.) x 150 mm long with
          a 100 ml reservoir.

    5.5   Stainless steel counting planchets, 2 inch diameter by 1/4 inch
          deep.

    5.6   Millipore filter apparatus, 47 mm.
                                     30

-------
6.  Reagents

    6.1   All chemicals should be of reagent grade, or equivalent, whenever
          they are commercially available.

    6.2   Ammonium hydroxide, 6N;  Mix 2 volumes 15f[ NH^OH (cone.) with 3
          volumes of water (carbonate-free).

    6.3   Anion exchange resin - Strongly basic, styrene, quaternary ammonium
          salt, 4% crossl inked, 100-200 mesh, chloride form (such as Dowex
          1x4, or equivalent).

    6.4   Ferric chloride carrier, 20 mg Fe+3/ml:   Dissolve 9.6 g of FeCl3.6
          H20 in 100 ml of 0.5N. HC1.

    6.5   Hydriodic acid:   HI (cone.), sp. gr. 1.5, 47%.

    6.6   Hydrochloric acid, 12N;  HC1 (cone.), sp. gr. 1.19, 37.2%.

    6.7   Hydrochloric acid, 8Nk  Mix 2 volumes 12f[ HC1 (cone.) with 1 volume
          of water.
    6.8   Hydrochloric acid, 6H_:  Mix 1 volume 12f^ HC1  (cone.) with 1 volume
          of water.

    6.9   Hydrochloric acid, 0.1N_:  Mix 1 volume 0.5f[ HC1  with 4 volumes of
          water.

    6.10  lodic acid, 1 mg/ml :   Dissolve 100 mg HIO  in 100 ml 4N. HN03.

    6.11  Nitric acid, 16^:  HN03 (cone.), sp. gr. 1.42,  70.4%

    6.12  Nitric acid, 4N_:  Mix 1 volume \6H_ HN03 (cone.)  with 3 volumes of
          water.

    6.13  Sodium hydrogen sulfite, 1% in HC1:  Dissolve 1  g NaHSO, in 100 ml
          6N. HC1 .

    Calibrations

    7.1   Determine a counting  efficiency (E), for known  amount of standard
          uranium (about 1000 dpm) evaporated from a 6-8  ml volume of a 1
          mg/ml HI03 solution in a 2-inch planchet.  If the standard solution
          is an HC1  solution,  then aliquot portions of that solution must be
          converted to nitrate/HN03 solutions, eliminating all chloride ions
          from the solutions.   This can be done by three  successive
          evaporations after adding 5 ml portions of 16H_ HN03 to aliquot
          portions of the standard in small  beakers (avoiding dry baking of
          the evaporated residue).  The final solutions of the standard
          aliquots are made by  adding 2 ml 4IN HN03 solution to the third
          evaporated residues.   Transfer the uranium standard aliquot
          solutions to 2 inch  diameter stainless steel  planchets.

                                     31

-------
          Complete the transfer by rinsing the beakers two times with 2 ml
          portions of 4N[ HN03 and evaporate to dryness.  Flame the planchets
          and count for, at least, 50 minutes for alpha particle activity.  A
          reagent blank should be run along with the standard aliquots and
          should be alpha counted.
                                                       A   B'
                          Efficiency, cpm/dpm, (E)  =    c             Eq- 1

               where:
               A  =  gross cpm for standard
               B1 =  cpm for instrument background
    7.2        C  =  dpm of standard used

          A uranium recovery factor R, is determined by the following
          procedure:  Spike one liter tap water samples with aliquots of
          uranium standard solution (500-1000 dpm per sample).  Take these
          spiked samples and a tap water blank through the entire procedure
          and count the separated and evaporated uranium for alpha particle
          activity.
                                                    (F   B}
                           Recovery factor, (R)  =  *  r  °'            Eq. 2
                                       C E
where:
C  =  dpm of uranium standard added
F  =  gross cpm of spiked sample
B  =  cpm of reagent blank
E  =  efficiency factor, cpm/dpm
8.  Procedure
    8.1   Measure the volume of approximately one liter of the water sample to
          be analyzed.

          8.1.2  If the sample has not been acidified,  add 5 ml  12N. HC1  and 1
                 ml ferric chloride carrier.

          8.1.3  Mix the sample completely and use pH paper to check the
                 hydrogen-ion concentration.  If the pH is >1, add 121^ HC1
                 until it reaches this value.

          8.1.4  Cover with a watch glass and heat the water sample to boiling
                 for 20 minutes.

          8.1.5  The pH must be checked again after boiling and if it is >1,
                 12H_ HC1 must be added to bring the pH back to 1.

          8.1.6  While the sample is still boil ing,gently add 6N. NH^OH to the
                 sample from a polyethylene squeeze bottle with the bottle
                 delivery tube inserted between the watch glass and the
                 pouring lip of the beaker.  The boiling action of the sample
                 provides sufficient stirring action.  Add 6f^ NHi+OH until
                 turbidity persists while boiling continues; then  add an

                                     32

-------
             additional  10 ml,  (estimated addition  from the squeeze
             bottle).

      8.1.7  Continue  to boil  the sample for 10 minutes more;  then  set  it
             aside for 30 minutes to cool and settle.

      8.1.8  After the sample  has settled sufficiently, decant and  filter
             the supernate through a 47 mm 0.45 micron  membrane filter,
             using the larger  millipore filtering apparatus.

      8.1.9  Slurry the remaining precipitate,  transfer to  the filtering
             apparatus and filter with suction.

      8.1.10 Place the filtering apparatus over a clean 250 ml  filtering
             flask, add 25 ml  8J\[ HC1 to dissolve the precipitate, and
             filter the solution.

      8.1.11 Add another 25 ml  8N^ HC1  to wash the filter, and  then  filter.

      8.1.12 Transfer  solution  to the 100 ml  reservoir  of the  ion exchange
             column.

      8.1.13 Rinse the side arm filtering flask twice with  25  ml  portions
             of SN^HCl.   Combine in the ion exchange reservoir.

8.2   Anion Exchange Separation

      8.2.1  Prepare the column by slurrying the anion  exchange resin with
             8N^ HC1 and pouring it onto a column about  13 mm inside
             diameter.  The height of the resin bed should  be  about 80
             mm.

      8.2.2  Pass the  sample solution through the anion exchange  resin
             column at a flow  rate not to exceed 5  ml/minute.

      8.2.3  After the sample  has passed through the column,  elute  the
             iron, (and plutonium if present),  with 6 column volumes of BN^
             HC1 containing 1  ml 47% HI per 9 ml of 8N.  HC1  (freshly
             prepared).

      8.2.4  Wash the  column with an additional  two column  volumes  of QH_
             HC1.

      8.2.5  Elute the uranium  with six column  volumes  of 0.11^ HC1.

      8.2.6  Evaporate the acid eluate to near dryness  and  convert  the
             residue salts to  nitrates by three successive  treatments with
             5 ml portions of  16f[HN03, evaporating to  near dryness each
             time.
      8.2.7  Dissolve the residue (may be very little  visible  residue)  in


                                 33
2 ml  4N_HN03.

-------
          8.2.8  Transfer the residue solution, using a Pasteur pipet, to a
                 marked planchet, and complete the transfer by rinsing the
                 sample beaker three times with 2 ml portions of 4N^ HN03.

          8.2.9  Evaporate the contents in the planchet to dryness, flame to
                 remove traces of HI03, cool, and count for alpha particle
                 activity.

    8.3   Column Regeneration

          8.3.1  Pass three column volumes of 1% NaHS03 in 6N. HC1 through the
                 column.

          8.3.2  Pass six column volumes of 61^ HC1 through the column.

          8.3.3  Pass three column volumes of water through the column.

          8.3.4  Pass six column volumes of 8N^ HC1 through the column to
                 equilibrate and ready the resin for the next set of samples.

9.  Calculations

                 Uranium alpha activity, pCi/1 = ^ J ? o°S°        Eq. 3
                                                  £.•£.£. X t K V

                 where:
                    S = gross cpm for sample
                    B = cpm of reagent blank
                    V = volume of sample used, ml
                    E = efficiency, cpm/dpm
                    R = recovery factor
                 2.22 = conversion factor for dpm/pCi

10.  Precision and Accuracy

     In a single laboratory test of this method, a stock uranium solution was
     prepared using tap water and spiked with an NBS uranium standard.  The
     calculated concentration was 26.7 pCi/1.  This stock solution was
     acidified with HC1 as a preservative.  Nine 1-liter aliquots were
     withdrawn and the procedure tested.  Individual results were 22.4, 22.5,
     24.0, 25.9, 26.9, 26.5, 24.6, 25.7 and 23.9 pCi/1.  The average
     concentration was 24.7 pCi/1 with a standard deviation of 1.7 pd'/l«
     From these data, the method shows a negative 7.4% bias and a precision of
     ±6.7% without the correction of the recovery factor.
                                     34

-------
                                  REFERENCES

1.   Bishop,  C. T., et al.   "Radiometric  Method  for the  Determination  of
     Uranium in Water",  EPA 600/7-79-093,  EMSL-LV, April  1979.

2.   Edwards, K. W.  "Isotopic Analysis of  Uranium  in  Natural Waters by Alpha
     Spectrometry", Radiochemical  Analysis  of Water, Geological  Survey Water -
     Supply Paper 1696-F,  U. S.  Government  Printing Office,  Washington, D.C.,
     1968.
                                     35

-------
                                            APPENDIX  C

                                  URANIUM IN DRINKING WATER
                                COLLABORATIVE TEST  - MAY,  1980
                                            DATA  SHEET
LABORATORY:

CONTACT  PERSON:

DATE:
 Sample
   ID
Vol
(ml)
Gross3
Counts
Counting
  Time
 (min.)
Counter
  Bkg
 (c/m)
  Bkg
Counting
  Time
(min.)
Uranium Alpha Activity15
       pC1/l
 (In sample or blank)
   1 a

   1 b

   1 c
   2 a

   2 b

   2 c
   3 a

   3 b

   3 c
   Blank a

   Blank b

   Blank c
   Recovery Factor (R), Determinations in %	

   Average of (R) 	%

   Counting Efficiency (E),  Determinations in %

   Average of (E) 	%
 aTotal counts  including background.

 bAs calculated from the equation in section 9 of the procedure.  For the purpose of this
   collaborative study, the reagent blank B is defined as the  cpm observed in one liter of your
   drinking water including the counter background.

 PLEASE SEND RESULTS TO C. A.  PHILLIPS, MONSANTO RESEARCH CORPORATION, MOUND FACILITY, MIAMISBURG,
 OH  45342, BY  JULY 18, 1980.  FTS 774-3228 (or 3927) or 513-865-3228 (or 3927)
                                             36

-------
                                 APPENDIX D

          COLLABORATIVE  STUDY  INSTRUCTIONS - URANIUM IN DRINKING WATER
                                  (May  1980)


1.   Use the procedure "Uranium  in Drinking Water - Radiochemical Method,
     Method 908.0"  dated May  1980.

2.   If you have not used a  procedure similar to the one enclosed, it would be
     advisable to analyze a  few  known samples, before analyzing the test
     samples.

3.   If you do not  have  the  ion  exchange resin, or other reagents needed for
     the analysis of the test  samples,  Mound Facility will furnish them if you
     desire.

4.   Use care in opening the  glass ampoules which contain the sample
     concentrates and are labeled  uranium concentrate 1, 2 and 3.  Each
     ampoule contains a  solution of  known concentration of uranium-238 +
     uranium-234 in 0.5M^ HC1.  The analyst should pipet exactly 5 ml of the
     concentrate and dilute this volume to one liter with drinking water in a
     volumetric flask.  Each  sample  should be prepared in triplicate from each
     uranium concentrate.

5.   The diluted one-liter samples just prepared are the samples to be
     analyzed.  For example,  the 5 ml aliquot of uranium concentrate #1
     diluted to one liter with drinking water is sample #1.  The analyst
     should therefore calculate  the  concentration of this diluted sample.

6.   The above samples are listed  as 1, 2, 3 on the enclosed Uranium In
     Drinking Water Collaborative Test  May 1980 Data Sheet.

7.   A uranium standard  of known concentration has been sent to you from the
     Environmental  Protection  Agency in Las Vegas.  This standard should be
     used to determine your  counting efficiency and uranium recovery factor.

8.   Additional amounts  of any uranium  concentrates are available if a sample
     is spilled, or an obviously incorrect uranium concentration is obtained,
     etc.

9.   Sample concentrations may be below 15 pCi/1.  Reagent blanks of your
     drinking water should be  made in your laboratory to be certain that they
     are not a problem at this uranium  concentration.
                                     37

-------
10.  The reagent blank  (B)  for the  purpose of this collaborative study is
     defined as the cpm observed  in one  liter of your drinking water including
     the counter background.   (Ordinarily, the  reagent blank is determined
     using distilled,  deionized water.)

11.  We are requesting  that uranium concentrations in all samples, counter
     efficiency determinations (E), recovery factors (R), and drinking water
     blanks be determined  in  triplicate.

12.  All samples should be  counted  for at least 50 minutes.  Longer counting
     times are advisable for  blanks and  any low-level sample.

13.  Individual counting data is  requested so that the counting statistics
     error can be resolved  from other errors.
                                     38

-------
                                  APPENDIX  E

                             QUESTIONNAIRE  SUMMARY
                     QUESTIONNAIRE ON  COLLABORATIVE STUDY
                    OF URANIUM IN DRINKING  WATER, MAY,  1980
Laboratory	Contact


1.  Did you make a blank correction?
                                                     Yes  19        No 	

2.  When you pipeted the tracer,  was  the volume
    of the pi pet calibrated?

                                                     Yes  16        No _3_

3.  Did you regenerate the anion  exchange resin?
4.  Did you deviate considerably  from any  part
    of the procedure?
                                                     Yes  16        No  3
                                                      Yes            No  19
    If so, what were these deviations?     (answer on other side)

5.  Do you believe that this would be a  good
    reference method for uranium in drinking  water?
                                                      Yes  15       No  2*
6.  Any other comments that  you  might  have would be appreciated.
*Two more were undecided.

                                     39

-------
                                  APPENDIX F

               COMMENTS FROM PARTICIPANTS IN COLLABORATIVE STUDY
1.   The general procedure appears to be a quick method for total  uranium
     determination.  The sample loss during handling appears to be minimal,
     and the analytical procedure appears much simpler than the fluorometric
     procedure.

2.   The procedure for calibration assumes that the addition of iodic acid to
     the standard duplicates the final  composition of the ion exchange eluate.
     The residue on a planchet after the ion exchange procedure was visibly
     less than the amount of iodic acid visible on a planchet prepared
     according to the calibration procedure.  Thus, the addition of iodic acid
     is not felt to be an appropriate method of correcting for self-absorption
     and counting efficiency simultaneously.  The counting efficiency
     determined by the calibration procedure is possibly too low and does not
     allow for any variability in the amount of residue.

3.   The quantitative transfer of the uranium eluant from the beaker to the
     planchet prior to counting is questionable.  In addition, "creeping" of
     the sample on the side of the planchet was observed thereby introducing
     some uncertainty in counting efficiency.

4.   The procedure should specify new anion exchange resin rather than
     regenerating resin.

5.   For every set of samples, analyze a spiked sample and one of  the samples
     in duplicate to check the recovery factor.

6.   The use of the term "column volume" would possibly be confusing to a
     technician.  In the entire procedure, column volume should be expressed
     as milliliters.

7.   A better estimate of the counting efficiency and a self-absorption factor
     can be made by determining the "weightless" counting efficiency and
     developing a self-absorption curve to use with the procedure.  The
     self-absorption curve can be made using iodic acid to increase the mass
     per unit area on the planchet as the same amount of uranium is added to
     each planchet.  The flaming procedure is not used as this weight must be
     accurately determined, and the flaming procedure could create variability
     in the planchet weight.

     The recovery factor would require weighing the planchet before and after
     the residue has been dried on it and using the net mass per unit area to
     find the self-absorption factor.

                                     40

-------
         The formula would be:



              «-w
         where:  F  =  gross  count per minute of spiked  sample
                 B  =  counts per minute of reagent blank
                 C  =  disintegrations per minute of  uranium standard added
                 E  =  counting  efficiency = cpm weightless std./dpm
                       weightless std.
                 A  =  self-absorption factor = cpm of std. of this mass/
                       unit/area/cpm of weightless std.

             Uranium alpha activity, pCi/1  =   2.2^'E^l A x°V°x R   Eq*  5

         where:  S  =  gross  cpm for sample
                 B  =  cpm of reagent
                 V  =  volume of sample used, ml
                 E  =  efficiency, cpm/dpm
                 A  =  self-absorption factor based on mass/unit area
                       (unitless)
              2.22  =  conversion factor for dpm/pCi
                 R  =  recovery  factor

8.   In step 8.2.6 by adding 16M  HN03 to a small volume of the evaporated
     uranium eluant, the iodate would be reduced to the free iodine; thereby,

     eliminating the iodate on the counting planchet.

9.   The procedure should call for the fresh'preparation  of NH[tOH for each
     analysis which would preclude the formation of the C03 ion.

10.  Uranium yields will vary  considerably depending on the amount of
     dissolved solids in the water sample.

11.  Some participants believed the method was good and easy to follow, while
     others believed the method was time consuming, complex, and would be too
     expensive for a routine method.

12.  This method should be an  alternative method of a fluorometric or alpha
     spectrometric method.
                                    41

-------