United States                    EPA-600 /R-92-141
           Environmental Protection
           A9encv                      July 1992    	
'EPA    Research and
          Development
          RECOMMENDED OPERATING
          PROCEDURE NO. 56:  COLLECTION
          OF GASEOUS GRAB SAMPLES
          FROM COMBUSTION SOURCES FOR
          NITROUS OXIDE MEASUREMENT
          Prepared for
          Office of Environmental Engineering
          and Technology Demonstration
          Prepared by
          Air and Energy Engineering Research
          Laboratory
          Research Triangle Park NC 27711

-------
                                EPA REVIEW NOTICE
        This report has been reviewed by the U.S. Environmental Protection Agency, and
        approved for publication. Approval does not signify that the contents necessarily
        reflect the views and  policy of the Agency, nor does mention of trade names or
        commercial products  constitute endorsement or recommendation for use.

        This document is available to the public through the National Technical Informa-
        tion Service, Springfield, Virginia 22161.
                                       DISCLAIMER



       This recommended operating procedure (ROP) has been prepared for the sole use of the Air

and Energy Engineering Research Laboratory (AEERL), U.S. Environmental Protection Agency,

Research Triangle Park, North Carolina, and may not be specifically applicable to the activities of other

organizations.

-------
                                         EPA-600/R-92-141
                                         July 1992
        RECOMMENDED OPERATING PROCEDURE NO. 56:

COLLECTION OF GASEOUS GRAB SAMPLES FROM COMBUSTION

         SOURCES FOR NITROUS OXIDE MEASUREMENT
                            Prepared by:
                          Jeffrey V. Ryan
                          Shawn A. Karns
                        Acurex Environmental
                        4915 Prospectus Drive
                          P.O. Box 13109
                   Research Triangle Park, NC 27709
                     EPA Contract No. 68-DO-0141
                     Technical Directive No. 92-066
                  EPA Project Officer: William P. Linak
                     Combustion Research Branch
              Air and Energy Engineering Research Laboratory
               Research Triangle Park, North Carolina 27711
                           Prepared for:

                  U.S. Environmental Protection Agency
                  Office of Research and Development
                       Washington, DC 20460

-------
                                   ACKNOWLEDGMENTS



       Assisting in the preparation of this procedure were Constance V. Wall and Shirley J. Wasson of



Research Triangle Institute, Research Triangle Park, NC, under EPA Contract 68-02-4291 for Quality



Assurance Support to AEERL.  Judith S. Ford, the QA Manager for AEERL, is responsible for the



development of operating procedures, including ROPs, for this EPA Laboratory.

-------
                                  TABLE OF CONTENTS

 Section                                                                            Page



 ACKNOWLEDGMENTS  	   ii

 1  INTRODUCTION	   1-1
   1.1  Scope and Application	   1-1
   1.2  Summary of Method  	   1-2
   1.3  Limitations	  	   1-2
       1.3.1   Limitations of Recommended Operating Procedures	   1-2
       1.3.2   Limitations of this Procedure	   1-3
   1.4  Definitions	   1-4

 2  START-UP	   2-1
   2.1  Personnel Requirements 	   2-1
   2.2  Facilities	   2-1
   2.3  Safety Precautions	   2-1
   2.4  Apparatus	   2-1

 3  PREPARATION OF SAMPLING  MATERIALS AND EQUIPMENT	   3-1
   3.1  Sample Container 	   3-1
       3.1.1  Sample Container Assembly	   3-1
       3.1.2  Sample Container Conditioning	   3-1
   3.2  SO2 Sorbent Cartridge	   3-3
       3.2.1  Sorbent Cartridge Preparation	   3-3
       3.2.2  Dispersion Tubes	   3-3
       3.2.3  SO2 Sorbent Preparation  	   3-4
       3.2.4  Filling SO2 Sorbent  Cartridge 	   3-4
   3.3  H2O Sorbent Cartridge	   3-5
       3.3.1  Cartridge Preparation	   3-5
       3.3.2  H2O Sorbent Preparation  	   3-6
       3.3.3  Filling H2O Sorbent  Cartridge 	   3-6

4  SAMPLING PROCEDURE	   4-1
   4.1  Summary of Procedure  	   4-1
   4.2  Sampling Procedure  	   4-2

5  REFERENCES	   5-1
                                            Hi

-------
                                     LIST OF FIGURES


  jure                                                                                Page

3-1.    Sample container schematic	  3-7

3-2.    SO2 sorbent cartridge assembly	3-8
       3-2A.  SO2 sorbent cartridge components	   3-8
       3-2B.  Dispersion tube hardware	  3-8
       3-2C.  Complete SO2 sorbent cartridge	  3-8

4-1.    N2O sampling system configuration	  4-4
       4-1A.  Location of sampling system	  4-4
       4-1B.  Sorbent/sample container schematic	  4-4

4-2.    N2O field data form	  4-5

4-3.    Sample custody sheet	  4-6
                                             iv

-------
                                                                                    Section 1
                                                                                   Revision 1
                                                                                   June 1992
                                                                                  Page 1 of 4
                                         SECTION 1

                                       INTRODUCTION



 1.1  SCOPE AND APPLICATION

        This procedure is applicable to the collection of gaseous grab samples from fossil fuel

 combustion sources for subsequent analysis of nitrous oxide (N2O).  This procedure details only the

 grab sampling methodology and associated equipment. Analysis of the collected grab samples is not

 within the scope of this procedure. A number of analytical procedures exist that are suitable for the

 measurement of N2O from grab samples.1'2

        Any gaseous stream resulting from the combustion of fossil fuels may be considered a post-

 combustion process stream. Gas streams from utility boilers, furnaces, and fluidized bed combustors,

 firing on natural gas, fuel oils, or coals are examples. This procedure has not been evaluated on waste

 incineration processes.

        The integrity of grab samples collected for N2O measurement has been demonstrated to be

 compromised under most common combustion sampling conditions.  A sampling artifact has been

 observed when nitrogen oxides (NOX), sulfur dioxide (SO2), and moisture (H2O), present in most

 combustion samples, react in the sampling containers to produce N2O.3  N2O generation in grab

 sample containers approaching 200 ppm has been observed.4

       Much of the data reported on N2O measurements from fossil fuel combustion sources were

obtained using grab sampling methods conducive to the sampling artifact.4-5 Because the previously

reported data are suspect, an economical method for assessing N2O emissions from fossil fuel

combustion sources was needed. As a result, this procedure was developed.  The goal of the


                                             1-1

-------
                                                                                    Section 1
                                                                                    Revision 1
                                                                                    June 1992
                                                                                  Page 2 of 4

 procedure is to minimize, but not eliminate, the N2O sampling artifact to levels acceptable for fossil fuel

 combustion source screening purposes. When this procedure is followed, the N2O generation within

 grab sample containers is less than 10 ppm in samples analyzed within 2 weeks of collection.

 1.2 SUMMARY OF METHOD

        This method is designed so that it can be used compatibly with continuous emission monitoring

 (OEM) sample delivery/conditioning systems or as a stand alone procedure.  Specifically, this method

 employs selective extraction to negate gaseous components to the point that the N2O generation in

 stored (1-2 weeks) sample containers is minimized to £ 10 ppm.  Sequentially, SO2 is neutralized and

 H2O removed from a flue gas sample stream before entering a Teflon®-lined stainless steel container.

 The neutralization and removal of SO2 requires the presence of H2O in the flue gas stream.  Therefore,

 the flue gas is collected before (upstream of) any moisture removal systems (e.g.,  refrigeration

 condensers, silica gel) that may be present in CEM flue gas conditioning systems.

        The flue gas sample is extracted using a vacuum pump.  The pump will  pull a flue gas sample

 from a post-combustion process stream and subsequently push the sample through the two-cartridge

 solid-sorbent system.  The first cartridge will neutralize and extract the SO2.  The second cartridge will

 extract the H2O. The SO2 cartridge must be placed before the H2O cartridge to ensure that a "wet*

 sample enters the SO2 cartridge. The flue gas sample will then flow into a grab sample container.  The

 grab sample container or "sample bomb" can then be sealed and stored for up to 2 weeks at  room

 temperature.  The sample containers can be analyzed for N2O at any point during  the 2-week holding

 period.

 1.3 LIMITATIONS

 1.3.1  Limitations of  Recommended Operating Procedures

        Recommended operating procedures (ROPs) describe non-routine or experimental research

operations where some judgement in application may be warranted.  ROPs may not be applicable to

activities conducted by other research groups, and should not be used in place of  standard operating

procedures or EPA-approved test methods. Use of ROPs must be accompanied by an understanding
                                             1-2

-------
                                                                                     Section 1
                                                                                    Revision 1
                                                                                    June 1992
                                                                                   Page 3 of 4

of their purpose and scope. Questions should be directed to the preparers or project personnel listed

in the Acknowledgments.

1.3.2 Limitations of this Procedure

       This procedure was developed to provide a method for evaluating the N2O emissions from

suspected high N2O emitting fossil fuel combustion sources. Researchers in the U.S. EPA's Air and

Energy Engineering Research Laboratory (AEERL) found that N2O generation in grab sample

containers, limited to less than 10 ppm, would be acceptable for the screening of fossil fuel combustion

sources.  This procedure may not be suitable for use on combustion sources where low concentrations

of N2O are anticipated.  For those  combustion sources, use of a grab sampling method where N2O

generation within grab sample containers has been demonstrated to be eliminated or use of an on-line

monitoring method is recommended.

       This procedure was developed for use with flue gases from conventional combustion sources

and processes.  Samples were designed to be collected upstream of any pollution control equipment or

on systems where pollution control equipment did not exist. This procedure may not be applicable for

use where sampling is performed downstream of pollution control devices or processes. This

procedure has been evaluated under both laboratory and field conditions.  Publication of these findings

is expected in the fall of 1992.

       During the development of the procedure, laboratory tests  were conducted to determine the  flue

gas NOX, SO2, and H2O concentration ranges where N2O generation would be consistently minimized

to less than 10 ppm over a 2 week period. This method was found suitable for use on combustion

systems with flue gas concentrations in the following ranges:

              SO2:          0 - 2500 ppm

              NOX:          0 -1000 ppm

              H2O:           5 - 25% (by volume)

       These flue gas concentration ranges were verified under actual field test conditions as well.

During these tests, the flue gas components of carbon monoxide (CO), carbon dioxide (CO2), and
                                             1-3

-------
                                                                                     Section 1
                                                                                    Revision 1
                                                                                    June 1992
                                                                                   Page 4 of 4

unbumed hydrocarbons, typically present in fossil fuel combustion process streams, were found to not

interfere with the sampling method performance.  Other common flue gas components such as

hydrogen chloride (HCI) and ammonia (NH3) were not evaluated and may act as interferences. Use of

this procedure where these compounds are present in the flue gas stream may not be applicable and

would require additional evaluation.

       Laboratory and field tests have found a two-cartridge sorbent system suitable for the collection

of three or four grab samples within a total operation time of 30  minutes. Any additional grab samples

taken with the sorbent system may result in the generation of  higher N2O concentrations due to SO2

breakthrough in the first cartridge.

1.4 DEFINITIONS

       Grab sample: a gaseous sample "grabbed" or pulled from a post-combustion process stream

       and stored in an inert, closed container.

       Sample container:  (sample cylinder, sample bomb) Tefk>n®-lined stainless steel container

       designed for safe containment of gas samples during drawing, storage, and handling or

       transport.

       Sorbent:  a substance that will hold or take up another substance (gases, solutes,  or

       liquids) with which they are in contact.

       Compression fitting: a compression type tube fitting that uses a nut and one or two ferrules

       to provide a leakproof, torque-free seal.
                                             1-4

-------
                                                                                   Section 2
                                                                                  Revision 1
                                                                                  June 1992
                                                                                 Page 1 of 3
                                         SECTION 2

                                         START-UP



 2.1  PERSONNEL REQUIREMENTS

       This procedure requires one technician trained on this ROP.  It is recommended that this

 technician be experienced in the sampling of gaseous emissions from combustion sources.

 2.2  FACILITIES

       This procedure requires a laboratory setup for the assembly of the two-cartridge solid-sorbent

 system, the space to include a fume hood, a source of deionized (Ol) water, and chemical storage.

 2.3  SAFETY PRECAUTIONS

       The cartridge assembly procedure uses calcium hydroxide [Ca(OH)2] and phosphorous

 pentoxide (P2O$)- Both substances are known to be hazardous if inhaled or swallowed. P2O5 and H2O

 produce a mild exothermic reaction.  The sample container conditioning procedure uses methanol

 (MeOH).  There are known hazards of fire and of poisoning due to ingestion. This procedure should be

 performed in a well-ventilated area.  Personnel protection should include safety glasses, laboratory

coats, and latex gloves.

2.4 APPARATUS

       Sample Container:

       •   600 cc stainless steel sample cylinder equipped with: Teflon® - coated inner surface, septum

          port,  1/4-in.* female nominal pipe thread (NPT) end connections
 For readers more familiar with the metric system, 1 in. - 2.54 cm.


                                            2-1

-------
                                                                               Section 2
                                                                              Revision 1
                                                                              June 1992
                                                                             Page 2 of 3
 •  Toggle valves (2), stainless steel, 1/4-in. male NPT x 1/4-in. compression fitting, Whitey Co.,
   (Cat. No. SS-92M4-S4) or equivalent
 •  Gas chromatograph septa, 3/8-in. outside diameter (OD),  Alltech Associates Inc., (Cat. No.
   6514) or equivalent
 •  Stainless steel washer, 8/32-in. OD
 •  1/4-in. stainless steel compression fitting nut (1) and end caps (2), Swagetock Co., (Cat. Nos.
   SS-402-1, SS-400-P, respectively) or equivalent
 •  Methanol, Burdick and Jackson, pesticide grade or equivalent
 •  Drying oven capable of 100 ±5 °C
 •  Nitrogen, zero grade
 •  1/4-in. Teflon® tape
 SO2 Sorbent Cartridge:
 •  120 cc refutable traps, gastight, Baxter Scientific, (Cat. No. MT120U-ATU) or equivalent
 •  Stainless steel O-ring-sealed straight-thread male connectors with 1/4-in. compression fittings
   (2), Swagetok Co., (Cat. No. SS-400-1-OR) or equivalent
 •  Teflon® tubing, 1/4-in. OD
 •  1/4-in. stainless steel compression fitting nut (4), Swagetock Co., (Cat. No. SS-402-1) or
   equivalent
•  1/4-in. stainless steel compression fitting ferrules (2 sets), Swagetock Co., (Cat. No. SS-400-
   SETS) or equivalent
•  1/4-in. ceramic-filled Teflon® compression fitting ferrules, Supelco, (Cat. No. 2-2086M) or
   equivalent
•  Calcium hydroxide, 98+ percent, ACS reagent grade, Aldrich Chemical Co. (Cat. No. 23,
   923-2) or equivalent
•  18-20 mesh (0.85-1.00 mm particle diameter) sand, Unimin Co. or equivalent
•  Large metal pan (e.g.,  13-in. baking pan)
                                       2-2

-------
                                                                           Section 2
                                                                          Revision 1
                                                                          June 1992
                                                                         Page 3 of 3
•  Play sand (optional)
•  U.S. Standard testing sieve -18 mesh (1.00 mm particle diameter), Fisher Scientific (Cat.
   No. 04-881 L) or equivalent (optional)
•  U.S. Standard testing sieve - 20 mesh (0.85 mm particle diameter), Fisher Scientific (Cat.
   No. 04-881M) or equivalent (optional)
•  Sieve shaker, Fisher Scientific (Cat. No. 04-909-4) or equivalent (optional)
H2O Sorbent Cartridge:
•  120 cc refutable traps, see SO2 sorbent cartridge apparatus
•  Stainless steel O-ring-sealed straight-thread connectors, see SO2 sorbent cartridge
   apparatus
•  Phosphorous pentoxide, indicating, Mallinckrodt Inc., (Aquasorb - Cat. No. 6063) or
   equivalent
•  Glass wool, Fisher Scientific (Cat. No. 11-390) or equivalent
Sampling Equipment:
•  Vacuum sample pump, air pumping rate of 8-12 L/min, Thomas Industries Inc., (Model No.
   107CA14TFEL-A) or equivalent
       [NOTE:  The sample pumps used during development of this
       procedure nominally delivered flow rates of 8-12 L/min under no-load
       conditions.  However, during actual sampling, the pressure drop
       across the sorbent cartridges restricted sample flow rates to 4-10
       Umin.  tt was under this flow rate range that the method was
       evaluated.  The sample pump flow rate requirement of £ 8 Umin was
       Identified as the minimum flow rate required to ensure adequate flow
       during sampling.]
•   Ftowmeter, 1-10 L/min range, Dwyer Instruments Inc., (Model RMA-21)
•   Teflon® tubing, 1/4-in. OD
                                     2-3

-------
                                                                                    Section 3
                                                                                   Revision 1
                                                                                   June 1992
                                                                                  Page 1 of 8
                                         SECTION 3

                  PREPARATION OF SAMPLING MATERIALS AND EQUIPMENT



 3.1  SAMPLE CONTAINER

 3.1.1 Sample Container Assembly

        Samples are collected in 600 cc stainless steel sample containers with their inner surface

 Teflon® coated.  The Teflon® coating provides a less reactant surface and eases sample container

 cleanup. The sample containers are equipped with toggle valves compatible with 1/4-in. compression

 fittings on both ends and a septum port for syringe sample extraction (Figure 3-1).  If the sample

 container is not equipped with an injection port, a 1/4-in. male connector compression fitting can be

 welded to the sample container. A 1/4-in. stainless steel compression fitting nut, containing an 8/32-in.

 washer, and a 3/8-in. OD gas chromatograph (GC) septum seal the port. This port allows direct

 extraction of the sample gas with a gaslight syringe from the sample container for injection onto a GC

 without contaminating the sample with room air.  For record keeping purposes, each sample container

 should be engraved or stamped with its own unique identification number.

 3.1.2 Sample Container Conditioning

       The sample containers  must be conditioned prior to use. The sample containers used for

sampling must be dean and dry. Any residual moisture within the sample container may cause N2O

generation in  excess of the limits of this procedure.  The sample containers are conditioned as follows:

       1. Remove nut, washer, and septum from side port.

       2. Remove end caps from toggle valves and open toggle valves.

       3. Soak sample containers in hot, soapy water for about 2 hours.


                                             3-1

-------
                                                                                    Sections
                                                                                   Revision 1
                                                                                   June 1992
                                                                                  Page 2 of 8
      4. Rinse three times with deionized (Dl) water and drain.
      5.  Rinse once with methanol.
      6.  Allow to dry.
              [CAUTION:  K oven temperature goes over 130 °C, the Teflon* coating
              and toggle valves may be damaged.]
      7.  Place sample containers in a hot oven (105 °C) for about 8 hours.
      8.  Replace washer and septum in the 1/4-in. nut.
      9.  Remove sample containers from oven, close toggle valves, replace end caps, and replace
          nuts and septa.
      10. Allow sample containers to cool.
      11. To purge and pressurize the sample containers, connect an H2O sorbent cartridge
          (Section 3.3) to the nitrogen tank regulator.  Attach a short piece of 1/4-in. OD Teflon®
          tubing to the cartridge, then affix a toggle valve to the free end of the tubing.  Next attach
          another piece of the Teflon® tubing to the free end of the valve.
      12. Remove end caps from a  sample container and attach container to the Teflon® tubing.
      13. Open the nitrogen cylinder (10 psi)*; then open the toggle valve after the H2O sorbent
          cartridge.
      14. Open the toggle valve on the sample container so  the container fills with nitrogen.
      15. Open the outlet valve on the sample container and remove the nut and septum from the
          injection port.
      16. Allow the sample container to purge for 1-2 minutes (at least 7 volume exchanges).
      17. Replace nut and septum.
      18. Close the outlet valve and then the inlet valve of the sample container.  The sample
          containers are now pressurized.  The pressure  within the sample containers will serve as a
For readers more familiar with the metric system, 1 psi - 6.9 kPa.
                                            3-2

-------
                                                                                        Sections
                                                                                       Revision 1
                                                                                       June 1992
                                                                                      Page 3 of 8

            quality control leak check prior to sampling use (see Section 4.2).

        19. Close toggle valve located after the H20 sorbent cartridge.

        20. Disconnect sample container and replace end caps.

 The sample containers are now clean, pressurized (~10 psi), and ready for sampling.

 3.2  SO2 SORBENT CARTRIDGE

 3.2.1  Sorbent Cartridge Preparation

        The sorbent cartridges used for SO2 removal are 120 cc refutable traps. The SO2 sorbent

 cartridge has the addition of two dispersion tubes to increase the reactive surface area.  The cartridges

 are fitted with stainless steel O-ring-sealed straight-thread connectors. The fittings for the SO2 sorbent

 cartridge are drilled out to allow 1/4-in. tubing to slide through them (17/64-in. drill bit).  (See Figure 3-

 2A.)

        The cleaning procedure for the SO2 sorbent cartridges is: rinse with water and use a soft bristle

 brush to remove grit, especially around the threads.  Rinse the cartridges with Dl water and dry.

 3.2.2  Dispersion Tubes

        To prevent channeling in the SO2 sorbent cartridge and to increase the reactive surface area,

 dispersion tubes are used. The dispersion tubes are designed to spray the flue gas through the solid

 sorbent thus maximizing the contact between the gas and sorbent.  Dispersion tubes are made from

 1/4-in. Teflon® tubing, 6 in. long. One end of the tubing is sealed with a 1/4-in. stainless steel

 compression fitting nut, ferrules, and cap (Figure 3-2B).  About 70 0.0225-in. holes (No. 74 drill bit) are

 drilled in a 2-in. section behind the nut and cap.  The collective area of the 70 holes is greater than the

 annular area of the Teflon® tubing to minimize flow restrictions. These tubes are  used at both the inlet

 and the outlet of the SO2 sorbent cartridge. The dispersion tubes slide through the drilled out O-ring-

 sealed straight-thread connectors.

       The dispersion tubes are secured in the cartridges using 1/4-in. ceramic-filled Teflon® ferrules.

These ferrules  do not become permanently compressed to the Teflon® tubing and allow for removal of

the dispersion tubes for easy breakdown and refilling of sorbent cartridges.
                                               3-3

-------
                                                                                     Sections
                                                                                    Revision 1
                                                                                    June 1992
                                                                                   Page 4 of 8
 3.2.3 SOg Sorbent Preparation
        The SO2 sorbent is a 10:1 ratio dry mixture of sand to Ca(OH)2.  This procedure requires the
 sand particles to be 18-20 mesh (particle diameter = 0.85-1.00 mm) in size.  This size range was found
 to maximize surface area with minimal pressure drop across the sorbent cartridge.  To prepare the SO2
 sorbent, weigh the desired amount of sand (e.g., 500 g) in a large pan, then weigh one-tenth of the
 sand amount of Ca(OH)2 (e.g., 50 g) and add to pan. Mix the sand and sorbent by hand for -15 min.
 Store the mixture in a desiccator.
        [NOTE:  If the 18-20 mesh sand Is not purchased, the following procedure
        may be  used (generic play sand Is suitable for this alternate procedure). This
        procedure requires an 18-mesh sieve, a 20-mesh sieve, and a sieve shaker.
        Put the  larger, 18-mesh sieve on top of the 20-mesh sieve. Pour the play
        sand Into the 18-mesh sieve. Place both sieves on the shaker.  Turn on
        shaker and let It run until sand is completely sifted. Collect the sand that
        goes through the 18-mesh sieve but does not go through the 20-mesh sieve.
        Discard  the larger and smaller particle-size sand fractions.]
 3.2.4 Filling SOg Sorbent Cartridge
        Before filling the sorbent cartridges, make sure the cartridges are clean and dry.  Fill the
 sorbent cartridges by packing the sand/Ca(OH)2 mixture around the dispersion tubes  located within the
 sorbent cartridge (Figure 3-2C). The dispersion tubes extend into the sorbent cartridge approximately
 3.5 in. from each end.  The dispersion tubes are held in place by the end caps of the sorbent cartridge.
 The dispersion tubes are secured to the end caps using a 1/4-in. compression fining nut and ceramic-
filled Teflon® ferrule, tightened to the O-ring seal connector compression fitting (Figures 3-2A and 3-
 2B).
       The following procedure details the filling/assembly of the SO2 sorbent cartridges:
       1.  Insert a dispersion tube into one of the sorbent cartridge end caps and attach with the
           ceramic-filled Teflon® ferrule.
                                             3-4

-------
                                                                                       Sections
                                                                                      Revision 1
                                                                                      June 1992
                                                                                     Page 5 of 8

        2.  Attach the body of the sorbent cartridge to the end cap (a gaslight seal is made against the

            O-ring located within the end cap). Adjust the dispersion tube so that ft extends into the

            sorbent cartridge approximately 3.5 in.

        3.  Holding the sorbent cartridge assembly vertically, add the sand/Ca(OH)2 mixture to the

            cartridge, filling ft to approximately 3.5 in. from the top. During the filling process, tap the

            cartridge occasionally to accelerate sorbent settling.

        4.  While holding the remaining dispersion tube upright inside the sorbent cartridge (the

            compression fitting on the dispersion tube should rest on the sorbent already within the

            cartridge), pack additional sorbent around the tube and fill the  sorbent cartridge to the top.

            Again, tap the sorbent cartridge occasionally while filling.

        5.  Slide the remaining end cap over the open end of the dispersion tube, and attach it to the

            sorbent cartridge. Make sure that surface of the sorbent cartridge contacting the O-ring

            seal in the end cap is smooth and free from sorbent mixture.

        6.  Secure the dispersion tube to the end cap with a ceramic-filled Teflon® ferrule and 1/4-in.

            compression fitting nut.

        The SO2 sorbent cartridge is attached to the pump and H2O  sorbent cartridge by a 1/4-in.

compression fitting nut and ceramic-filled Teflon® ferrule.

3.3  H2O  SORBENT CARTRIDGE

3.3.1 Cartridge Preparation

        The sorbent cartridges used for H2O removal are also the 120 cc refillable traps.  Dispersion

tubes are not used in the H2O sorbent cartridge.  Stainless steel, O-ring seal, straight-thread male

connectors with 1/4-in. compression fittings are used for attaching the H2O sorbent cartridge to the

associated sampling equipment.

        The cleaning procedure for the sorbent cartridges is: rinse with water and use a soft bristle

brush to remove grit, especially around the threads. Rinse the cartridges with Dl water and dry.
                                              3-5

-------
                                                                                     Sections
                                                                                    Revision 1
                                                                                    June 1992
                                                                                   Page 6 of 8
3.3.2  H2O Sorbent Preparation
       The  H2O is removed using commercial indicating P2O5. No sorbent preparation is needed.
3.3.3  Filling H2O Sorbent Cartridge
       The  H2O sorbent cartridges are filled with the indicating P2O5. The desiccant is held in place
with glass wool plugs at each end of the sorbent cartridge. To fill the sorbent cartridges, attach an end
cap to the cartridge body, hold it vertically, insert a glass wool plug in the capped end, and fill with -50
g of P2O5. Fill the remainder of sorbent tube with glass wool and attach the remaining end cap.  Make
sure that the surface of the sorbent cartridge contacting the O-ring seal in the end cap is smooth and
free from  sorbent particles.
                                              3-6

-------
                                                                            1/4-in. stainless steel nut
stainless steel
toggle valve
600 cc sample container
with septum port
                                                                            8/32-in. washer
                                                                           «*	
                                                                            3/8-in. OD septum
1/4-in. NPT  1/4-in. NPT
 (female)     (male)
                                                                      1/4-in. compression fitting
                                       Figure 3-1.  Sample container schematic.
                                                                                           JO «- 3J
                                                                                           $ § « S?
                                                                                           0 g <. 2
                                                                                           ^ *
                                                                                                                              oo ro -^ co

-------
CO
do
                          1/4-in. ceramic-filled Teflon ferrule
                                                                           120ccrefillabletrap
                     1/4-in. nu
                      m\>
         1/4-in. compression fitting
                                 7/16-20-in.
                                 straight
                                 thread
                                                                   stainless steel
                                                                   O-ring seal connector
                    1/4-in. ceramic-filled
                    Teflon ferrule
  Figure 3-2A.  SQ sorbent cartridge components.        end caP
                                                                            dispersion tube (w/0.0225-in. holes)  1/4.jn_ stain|ess stee| ferrules
O-ring seal    HHi end cap (O-ring inside end cap)

      Rgure 3-2B. Dispersion tube hardware.
                                                  Figure 3-2C. Complete SOj sorbent cartridge.
                                                   Figure 3-2. SO2 sorbent cartridge assembly.
                                                                                                                                             O) ^ Tl
                                                                                                                                             <5 c  to V)
                                                                                                                                             oo ro _»

-------
                                                                                   Section 4
                                                                                   Revision 1
                                                                                   June 1992
                                                                                 Page 1 of 6
                                         SECTION 4

                                   SAMPLING PROCEDURE



 4.1  SUMMARY OF PROCEDURE

        This procedure is used to collect gaseous grab samples from combustion sources for

 subsequent analysis of N2O. This procedure is designed so that it can be used compatibly with

 continuous emission monitoring (CEM) flue gas sample delivery/conditioning systems or as  a stand

 alone procedure.  When samples are obtained from a CEM flue gas delivery system, the N2O grab

 sample must be collected upstream of any moisture removing device(s); e.g., refrigeration and ice bath

 condensers or desiccants (Figure 4-1 A).

        The sampling system consists of a vacuum sample pump, a SO2 sorbent cartridge, a H2O

 sorbent cartridge,  and a grab sample container.  Important - Do not reverse order. The SO2

 cartridge requires  a wet sample.  Prepare the associated apparatus according to the procedures in

 Section 3.  In summary, a flue gas sample from a post-combustion process stream is drawn, under

 vacuum, by the sample pump. The sample pump, under positive pressure, forces the flue gas sample

 through the SO2 sorbent cartridge, H2O sorbent cartridge, and sample container consecutively (Figure

 4-1B).  Three separate, consecutive, grab samples can be taken per two-cartridge, solkj-sorbent,

 sample conditioning system. The three consecutive grab samples should be collected from the same

 sample location on each unit.  The sample containers are sealed and can be stored for up to 2 weeks

 at room temperature.  The containers may be analyzed at any point  during the 2-week holding period.

       As stated previously, this procedure was developed to characterize the N2O emissions from

different combustion sources.  To aid in this characterization of the combustion systems, certain


                                             4-1

-------
                                                                                    Sect ion 4
                                                                                   Revision 1
                                                                                   June 1992
                                                                                  Page 2 of 6

operating parameters of the combustion source need to be recorded. Figure 4-2, the N2O Field Data

Form, is an example of the information needed to characterize a combustion source.  Complete this

form during or immediately after the grab samples are collected.

       Also provide sample custody documentation (Figure 4-3).  The Sample Custody Sheet will aid

in the tracking of samples from the sampling location to the laboratory for the subsequent N2O analysis.

4.2 SAMPLING PROCEDURE

       To collect N2O grab samples:

       1.   Tee sample pump into a sample line with the inlet side of the sample pump connected to

           the tee (Figure 4-1A). Locate the pump before any CEM flue gas sample moisture

           conditioning systems.

       2.   Connect SO2 sorbent cartridge in series to outlet of sample pump (Figure 4-1B).  Connect

           H2O sorbent cartridge in series to the SO2 sorbent cartridge.

       3.   Remove 1/4-inch  end caps from sample container.

       4.   Connect sample container in series to H2O sorbent cartridge (Figure 4-1 B).

       5.   Attach flow meter to exit of sample container.

       6.   Leak check the sample container by opening the outlet valve of the sample container.  If

           the sample container is still pressurized (passing the teak check), the flow meter will

           momentarily indicate  flow.  If no flow is observed, replace the sample container.

       7.   Turn on sample pump.

       8.   Open the sample  inlet valve of the sample container.

       9.   Verify that sample flow exiting the sample container is between 4 and 10 L/min.

       10.  Purge the sample container for sufficient time (2-3 min) to allow  at least 7 volume

           exchanges to pass through the sample container.

       11.  Simultaneously close the sample outlet valve and sample inlet valve of the sample

          container.

       12. Turn off sample pump.
                                             4-2

-------
                                                                             Section 4
                                                                            Revision 1
                                                                            June 1992
                                                                          Page 3 of 6

13. Disconnect and remove sample container.

14. Replace both 1/4-in. end caps and tighten securely.

15. Repeat steps 3-14 for the two remaining sample containers.

16. Complete N2O Field Data Form (Figure 4-2). Identification numbers are stamped on the

   ends of the sample containers.

17. Complete Sample Custody Sheet (Figure 4-3).  Transfer samples to laboratory for

   subsequent N2O analysis.
                                      4-3

-------
                               stack
                                                    flow
To CEMs, dryart, ate
                                                      sample
                                                      pump

                                           Figure 4-1A. Location of sampling system.
from stack
            sample pump
                                                                                   Inlet
                                                                     HjOsorbent
                                          Rgure 4-1B. Sorbent/sample container schematic.
                                         Rgure 4-1.  N^O Sampling system configuration.
                                       outlet
                       sample container
 flow
meter
                                                                                                                                        o> ro -* *>.

-------
                                                                                  Section 4
                                                                                 Revision 1
                                                                                 June 1992
                                                                                Page 5 of 6
                             EPA/AEERL N2O FIELD DATA FORM

1. Sampler Identification -                       2. Untt Identification
Sampler's Name:

Company Name:
Company Address:
Phone:
Date Samples Taken:

Date Samples Sent: _
3. Fuel Characterization
Type of Fuel:
Fuel Analysis:
       Proximate Analysis (wt%)
        Volatile Matter:     	
        Fixed Carbon:     	
        Moisture:         	
        Ash:
       Ultimate Analysis (dry wt%)
        Carbon:          	
        Hydrogen:        	
        Nitrogen:         	
        Oxygen:          	
        Sulfur:           	
        Ash:             	
Heating Value (HHV):.
Btu/lb
5. Other Stack Emissions

  Oxygen (O2):

  Moisture
 Carbon Dioxide (CO2):
       N2O Cone.

 Sulfur Dioxide (SO2):

 Nitric Oxide (NO):

 Nitrogen Oxides (NOX):

 Carbon Monoxide (CO):
_ppm
_ppm
jppm
_ppm
                               Figure 4-2.
           Manufacturer of Unit:.

           Type of Unit:	
           Thermal Rating:

           Unit Name*:
                                                                             Btu/hr
                                         MW
            Unit Location*:
              'Optional (please include alphanumeric code if desired)
            4. Operating Conditions •

            Unit Load:	Btu/lb
            Stoichiometric Ratio:

            Excess Air	
            Type of Burners:
                                                NOX Control Used:.
            Flue Gas Cleaning Used:
                                                Operating Notes:.
            6.  N,O Analysis -

            Date Samples Received:.
            Date Analyses Performed:
                               Container f
  Analysis Performed By:
field data form.
                                         MW
                                             4-5

-------
                              N20 SAMPLE CUSTODY SHEET
                                                                                Section 4
                                                                               Revision 1
                                                                               June 1992
                                                                              Page 6 of 6
       Samples collected by

       Signature of collector

       Samples shipped by _
       Signature of transferor.

For Laboratory Use

       Samples received by _
       Signature of transferee
Date
Date
Date
Sample ID



Sample Contents \ Comments



Date Collected



AMf -. w>sX XvCwv$?W*<
FerLabifeg"
Date Analyzed;;
" tm%£\
V,
f • ' .''*$&
Comments \ Suggestions:
                                        Sample custody sheet.

                                            4-6

-------
                                                                                   Sections
                                                                                  Revision 1
                                                                                  June 1992
                                                                                 Page 1 of 1
                                        SECTION 5

                                       REFERENCES
1.     Ford, J.S., "Recommended Operating Procedure No. 45:  Analysis of Nitrous Oxide from
       Combustion Sources," 1990. EPA-600/8-90-053 (NTIS PB90-238502).

2.     Weiss, R.F., The Temporal and Spatial Distribution of Tropospheric Nitrous Oxide," Journal of
       Geophysical Research, 86, 7185-7195, 1981.

3.     Muzio, L.J., and Kramlich, J.C., "An Artifact in the Measurement of N2O from Combustion
       Sources," Geophysical Research Letters, 15:1369,1988.

4.     Linak, W.P., McSorley, J.A., Hall, R.E., Ryan,  J.V., Srivastava, R.K., Wendt, J.O.L., and Mereb,
       J.B., "Nitrous Oxide Emissions from Fossil Fuel Combustion," Journal of Geophysical Research,
       95(D6), 7533-7541, 1990.

5.     Ryan, J.V., and Srivastava, R.K., "EPA/IFP European Workshop on the Emission of Nitrous
       Oxide from Fossil Fuel Combustion," 1989. EPA-600/9-89-089 (NTIS PB90-126038).
                                            5-1

-------
                                 TECHNICAL REPORT DATA
                          (Please read Instructions on the reverse before completing)
1. REPORT NO.
 EPA-600/R-92-141
              3. RECIPIENT'S ACCESSION NO.
4. TITLE AND SUBTITLE
 Recommended Operating Procedure No.  56: Collection
  of Gaseous Grab Samples from Combustion Sources
  for Nitrous Oxide Measurement
              5. REPORT DATE
               July 1992
              6. PERFORMING ORGANIZATION CODE
7. AUTHOR(S)
 Jeffrey V. Ryan and Shawn A. Karns
              8. PERFORMING ORGANIZATION REPORT NO.
9. PERFORMING ORGANIZATION NAME AND ADDRESS
 Acurex Environmental
 P. O.  Box 13109
 Research Triangle Park, North Carolina  27709
              10. PROGRAM ELEMENT NO.
              11. CONTRACT/GRANT NO.
               68-DO-0141, Task 92-066
 12. SPONSORING AGENCY NAME AND ADDRESS
 EPA, Office of Research and Development
  Air and Energy Engineering Research Laboratory
  Research Triangle Park, North Carolina 27711
              13. TYPE OF REPORT AND PERIOD COVERED
               Task final; 1-12/91
              14. SPONSORING AGENCY CODE
               EPA/600/13
 15.SUPPLEMENTARY NOTES AEERL project officer is William P. Linak,  Mail Drop 65, 919/
 541-5792.
 16. ABSTRACT Tne Document is a recommended operating procedure (ROP),  prepared for
 use in research activities conducted by EPA's Air and Energy Engineering Research
 Laboratory (AEERL).   The procedure applies to the collection of gaseous grab sam-
 ples from fossil fuel combustion sources for subsequent analysis of nitrous oxide.
 The procedure details only the grab  sampling methodology and associated equipment.
17.
                              KEY WORDS AND DOCUMENT ANALYSIS
                 DESCRIPTORS
                                           b.lDENTIFIERS/OPEN ENDED TERMS
                           c. cos AT I Field/Group
 Pollution
 Sampling
 Nitrogen Oxide (N2O)
 Analyzing
 Fossil Fuels
 Combustion
  Pollution Control
  Stationary Sources
  Operating Procedures
13 B
14B
07B

21D
21B
18. DISTRIBUTION STATEMENT
 Release to Public
  19. SECURITY CLASS (ThisReport)
  Unclassified
                           21. NO. OF PAGES
    27
                                           20. SECURITY CLASS (Thispage)
                                           Unclassified
                                                                     22. PRICE
EPA Form 2220-1 (9-73)
5-2

-------