MIDWEST RESEARCH INSTITUTE
                                           I
REPORT
                           SOURCE TESTING REPORT
                          EMB PROJECT NO. 72-PC-12
                             UNION OIL COMPANY
                             Lemont, Illinois
                                    by

                               E. P.  Shea
                         Midwest Research Institute
                        Kansas City, Missouri  64110

                         EPA Contract No. 68-02-0228
                     (MRI Project No. 3585-C, Task No. 12)
MIDWEST RESEARCH INSTITUTE 425 VOLKER BOULEVARD, KANSAS CITY, MISSOURI 64110  •  AREA 816 561-0202

-------
                               SOURCE TESTING REPORT
                             EMB PROJECT NO. 72-PC-12
                                 UNION OIL COMPANY
                                 Lemont, Illinois
                                        by
                                    E. P. Shea
                            Midwest  Research  Institute
                           Kansas City, Missouri   64110

                            EPA Contract No.  68-02-0228
                         (MRI Project No. 3585-C, Task No.  12)
MIDWEST RESEARCH INSTITUTE 425 VOLKER BOULEVARD, KANSAS CITY, MISSOURI 64110 ° 816561-0202

-------
                                  PREFACE


          The work reported herein was conducted by Midwest Research

Institute (MRI) pursuant to a task order issued by the Environmental

Protection Agency (EPA) under the terms of EPA Contract No. 68-02-0228.

Mr. E. P. Shea served as the Project Chief and directed the MRI field

team consisting of:   Messrs., Gary Kelso, gas sampler; Fred Bergman, gas

chromatographic operator; Douglas Weatherman, methylene blue analyst;  and

Mike Serrone, iodometric titration analyst.  Mr. Winton Kelly,  EPA was

Project Officer and Mr. Richard Burr, EPA was the Field Engineer responsible

for collecting process data.  Mr. E. P. Shea prepared this final report.



Approved for:

MIDWEST RESEARCH INSTITUTE
Paul C. Constant, Jr.
Program Manager
26 January 1973
                                   ii

-------
                           I.   TABLE OF CONTENTS

                                                                       Page

II.    Introduction 	     2

III.   Summary of Results	     8

       A.  Observations 	     8
       B.  Suggested Method 	    15

IV.    Complete Gaseous Results 	    21

       A.  Gas Chromatograph	    22
       B.  lodometric Titration Procedure 	    36
       C.  Spectrophotometric  Method  	    41

V.     Sampling and Analytical Procedures 	    46

VI.    Process Operating Conditions 	    56

Appendix A - Determination of  H2S in Stack Gases	    57

Appendix B - Determination of  l^S in Refinery Fuel Gases	    62

Appendix C - Field Data Sheet	    69

Appendix D - Laboratory Report  	    81

Appendix E - Field Log	    98

Appendix F - Project Participants and Titles  	   100

-------
                             II.   INTRODUCTION


          Under the Clean Air Act of 1970,  as amended,  the Environmental

Protection Agency is charged with the establishment of  performance standards

for stationary source categories  which may  contribute significantly to  air

pollution.  A performance standard is a standard for emissions  of air pollu-

tants which reflects emission limitations attainable through the best emis-

sion reduction systems that have  been adequately demonstrated (taking into

account economic considerations).

          The development of realistic performance standards requires

accurate data on pollution emissions within the various source  categories.

Sampling and analytical techniques have to  be developed to acquire the  data.

A method for sampling and analysis of H2S is needed for the fuel gas systems

in the petroleum refining industry.  An analytical system for analysis  of

sulfur compounds using a gas chromatograph  equipped with a 36-foot by 0.085

inch polyphenyl ether and orthophosphoric acid Teflon column and a flame

photometric detector has been used by EPA.—   The objective of  this project

was to compare the chromatographic technique with a methylene blue-spectro-

photometric method and an  iodometric titration method at the Union Oil

Company Refinery in Lemont, Illinois.  This report presents the results of

the tests run at the refinery for EPA to determine:  (1) the applicability
_!/  J. D. Mulik, R. K. Stevens and R. Baumgardner, "An Analysis System
      Designed to Measure Multiple Maladrous Compounds Related to Kraft
      Mill Activities," TAPPI WATER and Air Conference, 4-7 April 1971.

-------
of the methylene blue-spectrophotometric method for analysis  of  t^S  in the





ranges of concentration present in the effluent from the amine scrubber,





(2) the applicability of the iodometric titration method,  (3)  the  compari-





son of the results from the wet methods with the results from the  gas





chromatograph and (4) changes in either or both procedures to make them





applicable to analysis of H2S in the concentrations present in the refinery





gas stream.




          Appendix A presents the EPA draft method for sampling  and analysis





of H2S by the methylene blue-spectrophotometric technique.  Appendix B




presents the EPA draft method for sampling and analysis of H£S using the




iodometric tiration.  These methods were tried, but had to be modified in




the field before any results could be obtained.  The mpdifications to the




methods are described in .this report.




          On Sundary, 25 June 1972, the equipment was shipped by truck to





Illinois.  Messrs. Kelso and Serrone drove the truck.  Messrs. Kelly and Burr




arrived at the site  Sunday afternoon.  On Monday, 26 June 1972, Messrs.




Shea, Bergman and Weatherman arrived at Lemont, Illinois,  and the  crew then




drove to the refinery arriving there about 10:15 a.m.  A meeting was




scheduled with the refinery personnel for 10:15, but had to be delayed




until 11:30, because of a truck fire across the street from the  refinery.




The process engineers who were members of Union Oil's fire brigade, were at





the fire.





          The equipment was installed Monday afternoon.  Figure  1  shows the





refinery fuel gas producing section with the MEA scrubbers and the fuel gas




mix tank.

-------
                                  UNION OIL COMPANY
                                  CHICAGO REFINERY
                                  FUEL GAS H2S REMOVAL

                                   H2S ABSORBERS (MEA) .
FUEL GAS
PRODUCERS
                            5-50 gr  H2S
                            per 100  scf
  Crude
  District
  Unit
           2 MM scfd
          10 MM scfd
                      Sour Gas
                      22.0 MM scfd
                                     100
                                     psig
                                    Lean
                                    Amine
       Control
       Valve
                                                   &
                          L«
                           L»-to
                                          Rich Amine
                                          to
                                          Regenerator
                                         3-30 gr  H2S
                                         per 100 scf
2.0 MM scfd
                      Sour Gas
                      2.5 MM scfd
                           100
                           psig
Lean
Amine
           1.5 MM scfd
                                            Rich Amine
                                            To
                                            Regenerator
              Fuel
              Gas
              Mix
              Drum
                                                               Fuel Gas
                                                           -^ to
                                                               Firing Line
                                                Natural Gas
                                                0.25grH2S
                                                                               Sample Point
                                         Figure 1

-------
          The sampling point was located after the control  valve between





the amine scrubber and the fuel gas mix drum.   This control valve regulates





the pressure to the mix drum at 50-60 psig from the 120-130 psig amine




scrubbing unit.  This is point A in Figure 1.   A Teflon lined pressure





regulator was installed after the 3/4 inch  valve in the low pressure line




as shown in Figure 2.  A 1/4-inch Teflon tube was attached  to the regulator





to serve as the sample line.  The tubing was enclosed in 3/8-inch copper




tubing for protection from accidental damage and steam tracing to keep the




temperature of the gas stream above ambient to prevent condensation.   The





sampling point for the impinger was 40 feet from the regulating valve.  A




Teflon tee was installed in the sample line with one branch for impinger




sampling and the other for GC sampling.  The GC was connected to the





impinger sampling tee with 10 feet of Teflon tubing.  In the morning, gas




flow was started in the sample line at least 20 minutes before a sample was




withdrawn to ensure that the line was purged.   The steam tracing was  started





at the same time.  A continuous flow was maintained in the  sample line until




sampling was finished for the day.  After the last sample was withdrawn




the gas and steam valves were closed.




          Ten gas samples were taken on Tuesday, 27 June 1972, using  the





midget impingers with Cd(OH)2 solution.  Two of the samples were ruined




because of excess pressure buildup in the sampling train, and one because





the 30 ml of Cd(OH)2 was exhausted.  Of the seven good samples four were




analyzed by the iodometric titration and three by methylene blue-spectro-




photometric method.

-------
                       UNION OIL COMPANY
                       CHICAGO REFINERY
                       FUEL  GAS H2S  REMOVAL
  Open
12 In. Pi
                     3/4 In. Valve    closed
                        High Pressure
                        Control Valve   Open
                  3 Ft -4 Ft
                     Y
                                                             3/4 In. Valve


                                                                X
Ground Level
                      Low Pressure
                                                                   6 Ft
               Sample Location A (Cat Cracker - Coker Stream)
                                Figure 2

-------
          On Wednesday there were 14 samples taken in the  Cd(OH)2  midget





impinger sampling train.  One sample was lost,  four were analysed  using the





methylene blue-spectrophotometric method and nine by the iodometric  titra-





tion.  Problems were encountered in the gas chromatograph  analysis all day





due to the increasing concentration of H2S saturating the  detector.





          There were 19 samples taken with the  impinger train on 29  June





1972, two samples were lost, four were analyzed with the methylene blue-




spectrophotometric, and 13 were analyzed using  the iodometric titration





method.  The final changes in the sample handling and analytical methods




were made this day.





          In all 24 analyses were made on the refinery gas stream  on Friday




30 June, using the chromatographic setup.  The  standard gas (46 ppm  H2S)




was injected into the chromatograph four times  to check standardization





before shutting down.  Since the standardization checked previous  results




very closely, the instrument was shut down and  the equipment packed  for





return to Kansas City.




          There were 24 samples collected in the Cd(OH)2 impinger train  on




Friday.  Six samples were analyzed using the modified methylene blue-




spectrophotometric method, and the other 18 using the modified iodometric





tritration method.  The first sample was taken at 8:53 a.m. and the  last




sample was finished at 5:52 p.m.  The equipment was packed and the truck





loaded for return to Kansas City.  We were finished at 8:00 p.m.,  and left





the refinery at this time.  On Saturday, two men drove the truck to  Kansas




City, while the other two returned by air to Kansas City.

-------
                         III.  SUMMARY OF RESULTS







A.  Observations





          The calibration curve for the methylene blue analysis prepared





in the laboratory before the field work was started was based on the  draft





method.  Because several changes were made in the methylene blue method





for analysis of H2S in the field, an attempt was made to establish a  stan-




dardization curve for the E^S levels obtained in the refinery gas stream.




It was not possible to obtain a standard curve for the methylene blue-





spectrophotometric determination of H2S at concentrations above 25 ppm.




The stream at the refinery contained from 200 to 500 ppm H2S.  Therefore,




the results using the methylene blue-spectrophotometric determination for




H2S were valueless for the refinery gas stream that was tested at Lemont.




Comparisons will not be made to the GC data because the concentration of





H2S cannot be calculated for the methylene blue analysis.




          Table I presents a comparison of the analytical results from the





flame photometric gas chromatograph analysis and the iodometric titration.




During the 4 days of sampling, there were 34 direct comparison analyses




between the iodometric titration and the GC.  Fourteen of these comparisons




were made while the iodometric titration was being modified and before the




final method for sampling and analysis was established.  On the 27th  there





was one comparison and the results are so far apart that it proved nothing.





The flame photometric detector was saturated and therefore the gas chromato-




graph data is not reliable.  The lower readings for H2S on 27 June are





                                    8

-------
                TABLE I
COMPARISON OF GC TO IODOMETRIC TITRATION


GC
1040
1050
1100
-
-
-
1126
1208
1335
1356
1417
-




0906
0918
0955
1003
1008
1014
1018
1023
1030
1035
1040
1045
1050
1055
1100
1119
1124
1129
1134
1139
1144
1149
1154
Time
lodometric
Titration


1111-1121
1337-1347
1501-1511
1525-1535

1156-1208

1402-1417
1424-1434
1439-1449
1523-1537
1549-1559
1601-1611
1710-1720

0925-0935








1042-1052


1057-1107

1117-1127

1130-1140





FOR H2S

DETERMINATION
Concentr at ion
Method Titration
EPA
a/


a/
a/
a/

a/

a/
a/
b/
b/
b/
y
Q /

D /








y
•

y

y

y





(ppm)


469
452
439
445
-
325

195
216
260
340
370
350
405

388








284


380

334

285






GC
(ppm)
262
273
261
-
-
-
362
405
423
449
375
-




430
546
498
488
437
470
281
415
454
462
515
491
479
450
485
475
491
498
525
516
499
495
495
                                                      Date

                                                     6/27/72
                                                     6/27/72
                                                     6/27/72
                                                     6/27/72
                                                     6/27/72
                                                     6/27/72
                                                     6/28/72
                                                     6/28/72

                                                     6/28/72
                                                     6/28/72
                                                     6/28/72
                                                     6/29/72

-------
  TABLE I (Concluded)
Time

GC
1241
1246
1251

1339
1334
1351
1413
1437
1448
1459
1000
1005
1030
1050
1115
1130
1155
1247
1300
1315
1330
1345
1415
1430
1447
1500
1517
1524
1617
1630
1645
1700
1715
1730
1745
1800
a/

b/

c/
d/
lodometric
Titration

1250-1300

1311-1321
1330-1340






958-1003

1027-1030
1050-1059
1119-1126


1252-1302


1322-1332

1412-1422
1427-1437



1521-1529
1613-1624
1630-1640
1643-1653
1658-1706
1715-1725
-
1739-1752
-
First modificatii
instead of Whai
Second modificat;
see Section V.
Final modificatii
See Section IV-A
Concentration
Method
EPA

£/

c/
£/





£/

c/
c/
£/


£/


£/

c/
£/



c/
c/
c/
c/
c/
£/

£/

Titration
(ppm)

544

560
519





398

355
345
312


411


437

442
402



335
394
293
396
360
373

406

GC
(ppm)
550
584
766

600
648
780
460
449
557
489
253d-/
354
321
347
368
378
392
440
472
448
476
400
412
403
478
500
640
378
366
293
335
330
448
372
446
431
                                             Date
                                            6/29/72

                                            6/29/72
                                            6/29/72
                                            6/29/72
                                            6/29/72
                                            6/29/72
                                            6/29/72
                                            6/29/72

                                            6/30/72
                                            6/30/72
                                            6/30/72
                                            6/30/72
                                            6/30/72
used fritted disc glass crucibles for filtering CdS
No. 40, see Section V.
 used 10% HCl to wash glassware instead of water,

see New Method, Section III-B.
23.
            10

-------
exactly opposite to the findings on the other dates.  On the 28th there





were three comparisons.  There was such a discrepancy between results that





validity of the iodometric titration procedure was in serious doubt.   On





the 29th the first 10 comparisons showed the same large spread between the





iodometric and the GC analyses.  Therefore, further changes were made in





the procedures for analyzing by the iodometric method.  The changes in the





iodometric titration procedure and the final method used are detailed in




Section IV.





          The last two comparison runs on the 29th and all of the comparison




runs on the 30th (with the possible exception of the first chromatograph




analysis) appear to be reliable comparison runs.




          Figures 3 and 4 are plots showing the results of analysis in




comparison of the values of the GC and iodometric titrations taken on the




29th and 30th of June.  Figure 3 is the comparison of the data for the





29th of June and shows the concentration of H2S in ppm on the ordinate and





the time at which the sample was taken on the abscissa.  The dots are results




of the GC analysis and the dashes are the iodometric analysis of the impinger




samples.  The dash corresponds to the sample time when the refinery stream




was being bubbled through the impinger train to collect the sulfide.   Upon




examination, Figure 3 shows that the concentration of H2S as determined by




the gas chromatograph was constantly changing.  The only iodometric titra-





tion samples that are shown on this figure are the ones using the final





modified method.  Examination of Figure 4 shows that the variation in







                                    11

-------
   800
   700
   600
 E
 Q.
 Q.

00


x  500
u
   400
   300




   250
       0906
                                                                                 • G. C. Sample

                                                                               _ Impinger Sample
0956
 1046




Figure 3
_

1116
1226
1316
1406
1456
                                                     TIME
                                               Comparison of  GC with  lodometric Titration

                                                 (29 June 1972 Start  0906)

-------
 E
 Q.
 Q.
CO
 CN
X
u
    500
450
    400
    350
    300
   225
                        — G. C. Sample
                         • Impinger Sample
                 I
                                                   _L
I
        0850   0940
                      1030      1120      1210      1300     1350
                                                 TIME - MlNUTES
        1440
1530
1620
1710      1800
                               Figure 4 -  Comparison of GC with lodometric Titration
                                            (30 June 1972)

-------
concentration during the sampling period was again extreme and the concentra-

tion of H2S in the refinery gas stream varied from about 200 to 500 ppm during

the  sampling period.  The concentration data obtained by the iodometric

titration procedure do not vary quite as greatly as the GC because the GC

is an instantaneous sample collected over a 1- to 3-second interval while

the impinger sample is collected during a 5- to 15-minute period.   This

would tend to average out the fluctuations that occur in this stream and

would tend to yield a flatter curve than the instantaneous sampling of the

GC.  During our sampling period not only did the concentration of  H2S vary

from time to time in the refinery gas but the pressure to our sample line

varied quite widely.  Instead of 50 to 65 psig supply,the pressure varied

from about 40 to 75 pounds at the point where we connected our pressure

regulator and sample delivery line.  This wide fluctuation in pressure put

a strain on the pressure reducing valve and required periodic attention and

a change of valve settings for proper delivery of gas to the sampling trains.

The sampling location was downstream of the control valve, thus changing

pressures could indicate changing flow rates of the gas stream.

          The variation in results is well within "t 107o in all cases and

there are a number of comparison readings that show essentially no difference

in concentration at all.  The conclusions that can be drawn from examination

of the data in Table I and Figures 3 and 4 are:  (1) that the iodometric

titration method as modified is a reliable method for on-the-spot  analysis

of refinery gas streams for hydrogen sulfide and, (2) the results  are well

within an average experimental error of 1 10%, using the gas chromatograph

data as reference.
                                     14

-------
B.  Suggested Method

          Soon after starting the sampling and analytical work at the re-

finery, it became apparent that the procedures, as written, were not adequate

for the sampling and analysis of I^S at the concentrations present in the

refinery stream.  Therefore, several modifications in the analytical and

sampling procedure were made.  For the determination of I^S in refinery fuel

gases by the iodometric titration, the recommended procedure based on our

work is:
    .  .  .  1.  Recommended Procedure of Determination of H2S in Refinery
                Fuel Gases

     1.   Principal and Applicability

          1.1  Principal

               H2S is collected in a series of midget impingers and

     reacted with alkaline cadmium hydroxide to form cadmium sulfide.

     The cadmium sulfide is dissolved in the impingers that it was

     collected in by adding an acidified iodine to the midget im-

     pingers.  A known volume of iodine solution is used with a

     known normality.  The excess iodine is titrated with thio-

     sulfate solution with starch as the indicator.

          1.2  Applicability

               This method should be applied only when specified

     by the test procedure for determining compliance with the

     new source performance standards.
                                    15

-------
2.  Interferences





3.  Apparatus





     3.1  Sampling





          The apparatus used shall be the same as that used in




Method 4, Federal Register. Volume 36, p. 24887, 23 December





1971, except that two dry impingers should be put into the





sampling train to make certain that no moisture is carried




into the dry gas meter.




     3.2  Analysis





          Glass stoppered 500 ml iodine no. flask for reaction




vessel.





4.  Reagents




     4.1  Sampling




          4.1.1  Absorbing solution - Mix 4.3 g cadmium sulfate





(3CdS04 8H20) and 0.3 g of NaOH in 1 liter of distilled water.




Mix well before using.





     4.2  Analysis




          4.2.1  Iodine Solution, 0.01 N





          Standardize daily as follows:  Pipette 100 ml of the 0.01 N





iodine solution into a 500-ml conical flask.  Titrate with the standard




0.01 N thiosulfate solution until the solution is a light yellow.  Add




a few drops of the starch solution and continue titrating until the





blue color just disappears.







                              16

-------
          4.2.2  Sodium Thiosulfate Solution.




          Sodium thiosulfate solution, standard 0.1 N.  For each





liter of solution, dissolve 24.8 g of sodium thiosulfate





(Na2S203.5H20) in distilled water and add 0.01 g of anhydrous




sodium carbonate (Na2 CO^) and 0.4 ml of chloroform (CHClg) to




stabilize.   Mix thoroughly by shaking or by aerating with nitro-





gen for approximately 15 minutes, and store in a glass-stoppered




glass bottle.




          Standardize frequently as follows:  Weigh into a 500-ml




volumetric flask about 2 g of potassium dichromate (K2Cr207)




weighed to the nearest milligram and dilute to the 500-ml mark.




Use dichromate which has been crystallized from distilled water





and oven-dried at 360°F to 390°F.  Dissolve approximately 3 g




of potassium iodide (KI) in 50 ml of distilled water in a glass-





stoppered, 500-ml conical flask, then add 5 ml of 20% hydro-




chloric acid solution.  Pipette 50 ml of the dichromate solu-




tion into this mixture.  Gently swirl the solution once and




allow it to stand in the dark for 5 minutes.  Dilute the solu-





tion with 100 ml to 200 ml of distilled water, washing down the




sides of the flask with part of the water.  Swirl the solution





slowly and titrate with the thiosulfate solution until the  solu-




tion is light yellow.  Add 4 ml of starch solution and continue




with a slow titration with the thiosulfate until the bright blue
                                17

-------
color has disappeared and only the pale green color of the





chromic ion remains.





          4.2.3  Sodium Thiosulfate Solution.





          Sodium thiosulfate solution, standard 0.01 N.  Dilute 100





- 0.01 ml of the standard 0.1 N thiosulfate solution in a volu-




metric flask to 1 liter with distilled water.




          4.2.4  Starch Indicator Solution.




          Suspend 10 g of soluble starch in 100 ml of distilled





water and add 15 g of potassium hydroxide pellets.  Stir until




dissolved, dilute with 900 ml of distilled water, and let stand




1 hour.  Neutralize the alkali with concentrated hydrochloric




acid, using an indicator paper similar to "alkacid" test ribbon,




then add 2 ml of glacial acetic acid as a preservative.





          Test for decomposition, by titrating 4 ml of starch





solution in 200 ml of distilled water, with the 0.01 N iodine





solution.  If more than 4 drops of the 0.01 N iodine solution




are required, make up a fresh starch solution.







5.  Procedure




     '5.1  Sampling




          5.1.1  Assemble the sampling train as shown in Figure 4-1.





Federal Register. Volume 36, p. 24887 connecting two midget




impingers with cadmium hydroxide, two dry midget impingers and




one midget impinger filled with silica gel in series.  Place






                               18

-------
15 ml of the collecting solution in each of the first two





irapingers.




          5.1.2  Purge the connecting line between the fuel





gas sampling valve and the first impinger.  Connect the sample





line to the train.  Read and record the initial reading on the





dry gas meter.




          5.1.3  Open the flow control valve and adjust the




sampling rate to read between 1.0 and 1.5 liters/minute.  Read




and record the meter temperature.





          5.1.4  Continue sampling for 10 minutes.  At the end




of this time, close the flow control valve and read the final




meter volume and temperature.




          5.1.5  Disconnect the impinger train from the sampling




line and cap the open ends.  Remove to the sample analytical





area.




     5.2  Sample Recovery and Analysis




          5.2.1  Combine the 50 ml of the 0.01 N iodine solu-




tion with 50 ml of 4 N hydrochloric acid in a 250-ml beaker.





Shake and carefully transfer the contents of the impingers




to a 500-ml iodine flask.  Rinse the impingers and connecting





glassware with 50 ml of the acidified iodine solution and care-




fully transfer to the iodine flask.  Replace stopper after each





additon of solution and shake.  Rinse the impingers and connect-




ing glassware with 25 ml of the acidified iodine solution and





                                19

-------
transfer to the iodine flask.  Rinse the impingers and connecting



glassware with the last 25-ml portion of the acidified iodine



and transfer to the iodine flask.  Then rinse the impingers,



glassware and 250-ml beaker with distilled water and quantatively



transfer to the iodine flask.  Stopper and shake well.  Analyze



within 30 minutes.



     5.3  Analysis



          5.3.1  Titrate the solution in the flask with 0.01



normal sodium thiosulfate solution, until the solution is light



yellow.  Add 4 ml of the starch solution and continue titrating



until the blue color just disappears.



          5.3.2  Run a blank determination with 30 ml of collect-



ing solution handled according to Section 5.2 to adjust the



value of the iodine consumption.





6.  Calculations



      Sample Calculations  of t^S by lodometric  Titration
           _ 0.2618 x [ANi  - (BN2-C)] g/scf
       H0S ' -
                          VMS
         C = ANB - BNB
             17.7 x Pm
       VMg = 	"TV x Vm cu ft
         a   (Tin + 460)
    0.2618 = (34 gr/mole H2S)  (0.0154 gr/mg)  (1,000 mg/g)

                        1,000  x 2 H2S equil
                               20

-------
           Calculations for Sample No. 24,  30  June 1972.



            VMS = 17.7 x 29,92 x 0.273 = Q>264 cu ffc


                     88 + 460




              C = (50 x 0.01)-(47 x 0.01) = 0.03



           C    ° 0.2618 x [50 x 0.01-(18.8 x 0.01 - 0.03)] =

            H2S                 0.264


                   0.254 gscf = 25.4 g/100 cf = 406 ppm





      Cu 0  =  Concentration  of  hydrogen sulfide in refinery  gas, gr/scf
       tlnO



        Pm  =  Meter pressure in.  of mercury  absolute




        Tm  =  Meter temperature °F




        Vm  =  Meter volume cu ft



         A  =  Volume  of standard  iodine used ml




        Ni  =  Normality of the  standard iodine  solution




         B  =  Volume  of standard  sodium thiosulfate used ml




        N2  =  Normality of the standard thiosulfate solutions




       VMS  =  Volume  of gas  sampled at standard conditions



       AN  =  Volume  of standard  iodine used x  the normality standard iodine
         "D

               used  in the  blank



       BNR  =  Volume  of standard  thiosulfate used x the normality  of the

               standard thiosulfate used in the blank






                       IV.   COMPLETE GASEOUS RESULTS






          This section contains  the results from the gas  chromatograph




analysis, the iodometric titration and the methylene blue-spectrophotometric



analytical  methods.
                                   21

-------
A.  Gas Chromatograph




          The results of the flame photometric gas chromatographic analysis





of the refinery gas stream are contained in Table II.   All of the data





recorded are presented in this table.




          There are a total of 60 analyses and all but two of them are





reliable readings based on our experience with the gas chromatograph and





the knowledge gained of refinery operations during this program.   We have




reported the time, the date, the gas chromatograph reading, the calibration




factor, and the concentration of H2S in ppm and grains/ 100 cubic feet.





          The GC was standardized using a 46 ppm standard H2S gas.  Because




of the high hydrocarbon content of the sample gas, the flame went out and




had to be relighted before the refinery gas could be analyzed by the





chromatograph.  Three samples of refinery gas were analyzed.  The chromato-




graph began malfunctioning and it was not possible to use the chromatograph





for analysis.  The nature of the trouble pointed to the electrometer.




          On June 28th, Mr. Bergman found and repaired a break in the power




cable from the electrometer to the detector head.  The chromatograph was




standardized with the 46 ppm H2S and five analyses were made.  The sample




loop for the preceding tests was 20 inches of 1/8-inch Teflon tubing.  The




200 to 500 ppm concentration of H2S in the refinery stream began to satur-





ate the detector so the sample loop was shortened by cutting 2- inch pieces




out of it starting at 1515.  The detector was still being saturated at 1720.





At 1730 the sample loop was shortened to 8 inches and standard l^S gas was
                                    22

-------
                                   TABLE II

                    CONCENTRATION OF H2S BY FLAME PHOTOMETRIC
                                                                        Gr/100 cf

                                                                          16.4
                                                                          17.1
                                                                          16.3

6/28/72         1126           452            0.80          362           22.6
                                                                          25.4
                                                                          26.4
                                                                          28.1
                                                                          23.4

6/29/72         0906           506            0.85          430           26.9
                                                                          33.2
                                                                          31.2
                                                                          30.5
                                                                          27.3
                                                                          29.4
                                                                          17.6
                                                                          26.0
                                                                          28.4
                                                                          28.8
                                                                          32.2
                                                                          30.7
                                                                          29.9
                                                                          28.2
                                                                          30.3
                                                                          29.5
                                                                          30.7
                                                                          31.2
                                                                          32.8
                                                                          32.2
                                                                          31.2
                                                                          31.0
                                                                          31.0
                                                                          34.4
                                                                          36.5
                                                                          47.9

                                                                          37.5
                                                                          40.5
                                                                          48.7
                                                                          29.7
                                                                          28.0
                                                                          34.8
                                                                          30.6
GAS CHROMATOGRAPHY
Time
1040
1050
1100
1126
1208
1335
1356
1417
0906
0918
0955
1003
1008
1014
1018
1023
1030
1035
1040
1045
1050
1055
1100
1119
1124
1129
1134
1139
1144
1149
1154
1241
1246
1251
1258
1339
1344
1351 ^
1413
1437
1448
1457

Reading
420
438
419
452
506
528
560
468
506
643
585
575
513
551
330
488
533
544
605
578
562
527
568
558
578
586
618
608
588
583
583
649
686
900
Saturated
705
763
918
540
528
655
575
23
Factor
0.625
0.625
0.625
0.80
0.80
0.80
0.80
0.80
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
0.85
Detector
0.85
0.85
0.85
0.85
0.85
0.85
0.85

ppm
262
273
261
362
405
423
449
375
430
546
498
488
437
470
281
415
454
462
515
491
479
450
485
475
491
498
525
516
499
495
495
550
584
766*
-
600
648
780
460
449
557
489


-------
                          TABLE  II  (Concluded)




             Time         Reading          Factor         ppm        Gr/100  cf
1000
1005
1030
1050
1115
1130
1142
1155
1247
1300
1315
1330
1345
1415
1430
1447
1500
1517
1524
1617
1630
1645
1700
1715
1730
1745
1800
177
250
227
245
260
280
290
326
618
700
664
727
591
609
595
708
740
730
432
418
335
383
377
511
425
510
492
1.43
1.43
1.43
1.43
1.43
1.43
1.43
1.43
0.675
0.675
0.675
0.675
0.675
0.675
0.675
0.675
0.675
0.675
0.875
0.875
0.875
0.875
0.875
0.875
0.875
0.875
0.875
253*
358
324
350
372
400
415
465
440
472
448
476
400
412
403
478
500
492
378
366
293
335
330
448
372
446
431
15.8
22.4
20.2
21.9
23.3
25.0
25.9
29.0
27.5
29.5
28.0
29.8
25.0
25.8
25.2
29.8
31.2
30.8
23.6
22.9
18.3
20.9
20.6
28.0
23.2
27.9
26.9
These results are questionable.
                                   24

-------
passed through the chromatograph several times to help clean out the detec-





tor.  The chromatograph was shut down and all sampling discontinued for





the day.  There were seven samples passed through the GC from 1500 to 1720,





but since the detector was saturated on all samples no attempt was made to





calculate the concentration of H2S.





          Thursday morning, 29 June 1972, the sample loop on the chromato-





graph was shortened to a volume of 0.1 ml before analysis began.  Standard




HoS (46 ppm) was passed through the chromatograph for standardization.





Sampling began at 9:06 a.m.  Three analyses were made to scan for components




other than I^S, CH.,SH (methyl mercaptan) was the only other component de-




tected in more than trace quantities.  At 9:55 it was decided to inject gas




from the sample line into the instrument every 5 minutes, 13 analyses were




made in this manner.  At 11:00 a.m. a 19-minute run was made on one injec-




tion.  The only components found were H2S and CH3SH.  Nineteen more analyses




were made with 5 minutes between injections.  At 3:00 p.m. the chromatograph




again started malfunctioning.  The indications were that there was no H2S





in the effluent.  The standard (46 ppm) H2S was injected into the chromato-





graph.  The trace showed a blip instead of a peak.  Several more injections





were made at 10-second intervals with the standard gas, and finally, a peak




about one-fourth of the proper size, did show up.  It was theorized that the




column had become saturated with the amine (MEA) from the scrubber system.




A 1-foot section of the column was cut off and the column reversed.  Then





several more injections of standard gas were made at 10-second intervals.




A peak about one-half of the normal size was obtained.





                                     25

-------
          At 6:00 p.m., it was decided to bake the column over night.





The temperature was raised from 60°C to 110°C, the nitrogen carrier gas





was left on and the rest of the instrumentation turned off.





          Friday morning the temperature of the column was reduced to 60°F,





the instrumentation turned on and the flame lit.  After the chromatograph





stabilized, standard (46 ppm) HoS was injected into the column several times





for standardization.  Analysis of the refinery stream started at 10:50 a.m.




and there were five analyses before lunch.  Between 12:47 p.m. and 4:00 p.m.,




11 analyses were made.   At 4:00 p.m., four injections, at 4-minute intervals,




were made with the standard (46 ppm) I^S to check standardization.  Then




eight more injections were made and analyses recorded.





          Table III shows the sample calculations for determining the




hydrogen sulfide concentration from the GC analysis.





          The following week after the field test, the chromatograph was




reassembled in Kansas City and standard  gases  run through it to obtain





calibration for the various sulfur compounds.  Forty-six parts per million




hydrogen sulfide gas was injected and the response was the same as obtained




during our work at Lemont, Illinois, the previous week.  In addition, a




mixture of the gases, carbonyl sulfide, methyl mercaptan, and sulfur dioxide




was injected to obtain calibrations to determine if any of them were present




in the refinery gas tested at Lemont.  Separate calibrations were also ob-




tained for carbon disulfide and ethyl mercaptan.  Then a mixture of nitro-





gen, COo and S07 was injected to determine if C02 and S0? were present in








                                    26

-------
                                TABLE III

            SAMPLE CALCULATIONS FOR HYDROGEN SULFIDE CONCENTRATION

 1.  Hydrogen Sulfide Concentration from Gas Chromatograph Analysis
              R x 64 x 46 ppm
                  Rs x 4
            Calculation for 1,800 hr on 30 June 1972


                 C    = 492 x 64 x 46 = 4n
                  H2S       840 x 4         PPm


       CH s = Concentration of H2S in the refinery gas, ppm.

          R = The readings on the recorder chart from the chromatograph when
                the refinery gas is analyzed.

         Rs = The reading on the recording chart when standard I^S is analyzed
the refinery gas.  After examining the calibration curves and the chromato-

graph recordings from the work at Lemont, Illinois, it was determined that

the only materials found by chromatography were H2S,  methyl mercaptan and

a trace of carbon dioxide.  Table IV shows the relative retention (a) to

the internal standard of methane for the components which were expected to

be in the refinery gas that was analyzed.  Table V shows the components that

were resolved when the large 20-inch sampling loop was installed in the

chromatograph.  These data were taken on Tuesday, June 27th, and show that

there was a trace of methane, ethane, a concentration of H2S of about 273

ppm, five unknown hydrocarbons with the relative retention of 2.06, 2.41,

2.64, 2.92, and 3.58.  Methyl mercaptan and carbon disulfide at concentra-

tions of about 9.4 ppm and 0.9 ppm, respectively, were also found.

                                    27

-------
                                  TABLE IV
           RELATIVE RETENTION n, TO INTERNAL STANDARD OF METHANE
       Component

          CH4
          C2H6
          C02
          C3Hg
          H2S
          COS
          S02
          CH3SH
          CS2
          C2H5SH
 Retention
   Time-/

   14.8
   16
» 17
   18
   23
   25
   45.5
   64
   120
   167
                                        Relative
                                        Retention
                                            .0
                                            ,08
                                            .15
                                            ,22
                                            ,55
                                            ,69
                                            ,07
                                          4.32
                                          8.11
                                         11.28
a/  Measured in millimeters on recorder chart L&N No. 492000, with recorder
      chart speed of 1 inch per minute.

                                   TABLE V

                       COMPONENTS  RESOLVED USING LARGE
                       SAMPLING  LOOP ON 27 JUNE 1972
                     (Samples  Taken at 10:40 and 12:08")
   a
 1.0
   08
   55
   06
 2.41
 2.64
 2.92
 3.58
 4.32
 8.11
      Compound

        CH4
        C2H6
        H2S
Unknown Hydrocarbon
Unknown Hydrocarbon
Unknown Hydrocarbon
Unknown Hydrocarbon
Unknown Hydrocarbon
        CH3SH
          2
                   Approximate Concentration (ppm)
                                273
                                  9.4
                                  0.09
Note:  See Figure 5 page 32 for peaks attributed to these compounds.
                                     28

-------
Table VI contains the report on the concentration of methyl mercaptan found

in the refinery stream on June 28th, 29th and 30th.  The data are reported

in parts per million with the time and date at which it was found.


                                 TABLE VI

           CONCENTRATION OF METHYL MERCAPTAN IN REFINERY STREAM
 Time

 11:26
  9:06
  9:18
  9:55
 10:03
 11:54
 10:00
 10:05
 10:50
 11:15
 11:30
 11:42
 11:55
 12:47
 13:00
 13:15
 13:30
 13:45
 14:15
 14:30
 14:47
 15:00
 15:17
 15:24
 16:17
 16:30
 16:45
 17:00
 17:15
 17:30
 17:45
 18:00
(28-30 June 1972)

Date
6/28/72
6/29/72
6/29/72
6/29/72
6/29/72
6/29/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72
6/30/72

Methyl Mercaptan
(ppm)
7.5
6.0
13.3
17.5
10.2
13.6
3.4
6.8
2.6
3.4
8.5
8.3
10.9
11.1
11.2
12.3
11.7
11.2
10.7
12.8
12.5
13.3
14.2
15.6
9.9
9.6
11.1
12.8
15.4
14.3
16.7
12.8
                                    29

-------
          Figures 5 through 9 are reproductions of the recorder chart from

the gas chromatographic analysis for H2S and other components in refinery

gas.  Figure 5 shows the first determinations on the 27th of June and also

shows the line voltage fluctuation that gave us trouble that morning and

blew several fuses in the chromatograph as well as the spectrophotometer.

Examination of the figure shows the components that can be identified in

the refinery gas stream.  These are hydrogen sulfide, methane,  ethane,

methyl mercaptan, five unknown hydrocarbons and carbon disulfide.

          Figure 6 is the recorder trace for the chromatograph and shows

the standardization of the instrumentation as well as a trace for E^S and

methyl mercaptan.  Figure 7 is a reproduction of the recorder trace for the

afternoon of the 28th.  This was the day that the detector was becoming

saturated because of the high concentration of hydrogen sulfide in the

refinery gas stream.  During this period the sample loop was being shortened

so that an analysis could be made using the gas chromatograph.  The first

trace shows a rounded point rather than a sharp, well-defined point

indicating that the photo detector was still being saturated.  Both hydrogen

sulfide and methyl mercaptan can be seen in the trace on this day.  Figure

8 shows a trace from the 29th of June and shows the clean spectrum that

should be obtained using the detector.  Also, it shows the period when a

sample was injected approximately every 5 minutes.  The only two components

being resolved were t^S and CH3SH, with the possible exception of a slight

trace of CC^.  Carbon dioxide would give a negative peak in respect to

and methyl mercaptan.  Figure 9 is a trace of the 30th of June and shows

the type of recording and detection experienced at the time of shutting

down at 1800 hours on 30 June.
                                     30

-------
Figure 5 - GC Trace
         31

-------
70 0
        71
                  It**
             Figure 6 - GC Trace
                  32

-------
Figure 7 - GC Trace



         33

-------
 $00
   ?*
/CO
             Figure 8 - GC Trace


                   34

-------
Figure 9 - GC Trace
                                       •" A
          35

-------
          Figure 10 is a calibration chart for hydrogen sulfide.   A standard

gas (46 ppm hydrogen sulfide) was introduced into the gas chromatograph four

different times and the readings obtained were used to determine  the multi-

plying factor for converting from scale reading to concentration  of hydrogen

sulfide in the refinery gas.  Figure 11 is a copy of the calibration curve

for methyl mercaptan, carbon dioxide, carbonyl sulfide, and sulfur dioxide.

One thing that becomes apparent in the examination of this calibration curve,

is that carbonyl sulfide and hydrogen sulfide are eluted from the column at

nearly the same time and therefore, it would be difficult to separate and

quantify carbonyl sulfide and hydrogen sulfide with the 36-foot Teflon

column used at Lemont, Illinois, Union Oil Refinery.*  It was from calibra-r

tion curves of this type that the value of methyl mercaptan and the other

constituents were determined so that the readings on the chart could be

converted into ppm.


B.  lodometric Titration Procedure

          Table VII contains the results of the iodometric titration for

determination of hydrogen sulfide in the refinery gas stream from the amine

scrubber.
*  The carbonyl sulfide, if present, would show a double peak with hydrogen
     sulfide or a peak on the back side of the hydrogen sulfide peak.
     Examination of the chromatograph trace on June 27th and 28th (Figures
     5 and 6), when the large sample loop was used, does not show any
     indication of the carbonyl sulfide peak.  Therefore,  the concentra-
     tion of carbonyl sulfide is less than 0.5% of the concentration of
     hydrogen sulfide in this refinery gas stream.
                                    36

-------
Figure 10 - Standardization Using 46 ppm H0S
                        37

-------
              cos
Figure 11 - Standardization Using Gas Mixture
                     38

-------
              TABLE VII
CONCENTRATION OF H2S DETERMINED BY
IODOMETRIC TITRATION
Sample No.
and Time
2
4
6
7
1
3
4
5
8
9
10
13
1
4
5
6
7
8
9
10
14
15
16
17
18
1
2
3
4
5
6
7
9
11
13
14
17
19
20
21
22
23
24
1042-1052
1337-1347
1501-1511
1525-1535
1156-1208
1402-1417
1424-1434
1439-1449
1523-1537
1549-1559
1601-1611
1710-1720
0925-0935
1042-1052
1057-1107
1117-1127
1130-1140
1250-1300
1311-1321
1330-1340
1523-1533
1539-1550
1554-1607
1612-1626
1632-1642
0853-0908
0929-0939
0958-1003
1027-1030
1034-1044
1050-1059
1119-1126
1252-1302
1322-1332
1412-1422
1427-1437
1521-1529
1613-1624
1630-1640
1643-1653
1658-1706
1715-1725
1739-1752
VM
(cu ft)
0.318
0.323
0.323
0.318
0.323
0.517
0.565
0.393
0.301
0.325
0.369
0.322
0.335
0.358
0.333
0.394
0.413
0.335
0.307
0.388
0.381
0.300
0.303
0.302
0.521
0.304
0.364
0.262
0.251
0.578
0.349
0.368
0.423
0.316
0.309
0.298
0.354
0.034
0.365
0.359
0.377
0.347
0.273
TM
(°F)
80
84.5
84.5
84.5
74
72
71.5
71.5
72
73
72.5
71
68.5
70
71
71
70
74.5
75
74.5
74.5
75.5
75.5
76
75.5
75
79
84.5
91
88
87
85.5
87
87
87.5
89
88
88
88
88
88
88
88 ,c
VMS
(cu ft)
0.312
0.315
0.315
0.310
0.322
0.516
0.563
0.391
0.299
0.323
0.367
0.321
0.336
0.358
0.332
0.393
0.413
0.333
0.304
0.386
0.379
0.296
0.299
0.298
0.515
0.300
0.358
0.255
0.241
0.559
0.338
0.357
0.410
0.306
0.299
0.288
0.342
0.294
0.353
0.338
0.364
0.336
j 0.264

Gr/100 cf
29.3
28.2
27.4
27.8
20.3
12.2
13.5
16.2
21.2
23.1
21.8
25.3
24.2
17.2
23.7
20.9
17.8
34.0
35.0
32.4
24.1
23.8
20.3
19.4
17.0
18.4
21.4
24.9
22.2
15.4
21.6
19.5
25.7
27.3
27.6
25.1
20.9
24.6
18.3
24.8
22.5
23.3
25.4

ppm
469
452
439
445
325
195
216
260
340
370
350
405
388
284
380
334
285
544
560
519
386
380
324
310
272
294
342
398
355
246
345
312
411
437
442
402
335
394
293
398
360
373
406

Date
6/27/72



6/28/72







6/29/72












6/30/72


















-------
The data are presented as sample number, time of sampling, date sampled,




the meter volume, meter temperature of the dry gas meter in the sampling



train, the calculated volume of gas samples at standard conditions, and



the concentration of H2S in grains/100 cubic feet and ppm.  In all there



were 43 iodometric titrations.  The titrations on Tuesday the 27th and



Wednesday the 28th are of very little value as it is certain that the method



of sampling, handling and analysis was allowing the escape of hydrogen sul-



fide when the cadmium sulfide was dissolved in hydrochloric acid.  Samples



1 through 7 on Thursday the 29th were also analyzed before the final ana-



lytical and sampling procedures were established and these values are of



questionable value.  Samples Nos. 8 through 18 on Thursday the 29th and all



of the samples on Friday the 30th were collected and analyzed by the modified



procedure which is reported in Section III-B of this report.





           Concentration of  Hydrogen Sulfide by  Iodometric  Titration


           = 0.2618 x [ANj_ - (BN2-C)]  g/scf


       V"              v^




         C  = ANB -  BNB



            17.7 x Pm   „      -
      VMq  = 	;	 x Vm cu ft
        S   (Tm +  460)



   0.2618  = (34 gr/mole H2S>  (0-0154  gr/mg)  (1,000  mg/g)


                       1,000 x 2 H2S equil




           Calculations for  Sample  No.  24,  30  June  1972
                                    40

-------
        VM    17.7 x 29.92 x 0.273   . „,.
        VMS - 	 x 0.264 cu ft

                  88 + 460
          C =  (50 x 0.01)-(47 x 0.01) = 0.03




            _  0.2618 x [50 x Oo01-(18.8 x 0.01 - 0.03)]
       C


        V             0.264
H0S
                     0.254 gscf = 25.4 g/100 cf = 406 ppm





      CH Q = Concentration of hydrogen sulfide in refinery gas, gr/scf
       tlrt O




        Pm = Meter pressure in. of mercury absolute





        Tm = Meter temperature °F





        Vm = Meter volume cu ft





         A = Volume of standard iodine used ml





        N, = Normality of the standard iodine solution





         B = Volume of standard sodium thiosulfate used ml





        N2 = Normality of the standard thiosulfate solutions





       VMg = Volume of gas sampled at standard conditions





       ANg = Volume of standard iodine used x the normality standard iodine


               used in the blank





       BN  = Volume of standard thiosulfate used x the normality of the


               standard thiosulfate used in the blank.







C.  Spectrophotometric Method





          The modified method for methylene blue-spectrophotometric deter-





mination of hydrogen sulfide in the refinery gas  stream was used in trying





to draw a standard curve so that  the absorbance data  obtained  in the sampling





at the Union Oil Refinery at Lemont,  Illinois,  could  be related to  the  con-





centration of H2S in ppm.   An attempt was made  to prepare  a calibration






                                     41

-------
curve for 5, 10, 25, and 50 ppm H2S.   Figure 12 shows that at the values

below 10 ppm, the curve has the right slope, but is straightened out  above

10 ppm and went backwards at all values above 25 ppm.  It was not possible

to get an absorbance reading above 35 ppm H^S.

          Because several changes were made in the methylene blue-spectro-

photometric method for analysis for H2S in the field, a new standardization

curve for the methylene blue method was needed.

          The curve, Figure 13, shows the results of making up a 200  ppm,

350 ppm, and a 500 ppm standard using analyzed reagent grade cadmium  sul-

fide.  The curve has a negative slope, or in other words, higher parts  per

million H2S gave a lower absorbance reading.  This same type of response

was found in the calibration work for a standard curve before going to  the

refinery.  It was then decided to see if mixing the hydrochloric acid and

the ferric chloride before dissolving the cadmium sulfide would allow prep-

aration of a calibration curve in the range of 200 to 500 ppm hydrogen  sulfide.

As shown in Figure 14, the curve does have a less steep slope; however, it

is still negative and is not useful for determining the concentration of

H2S in the refinery stream at the concentrations found in the Union Oil

Refinery at Lemont.  While dissolving the cadmium sulfide in the acid

solution, the smell of H2S escaping from the flask was noticeable. The

higher the concentration the more t^S escaped while dissolving the cadmium

sulfide.  The ability to analyze the refinery stream for l^S using the

methylene blue method is entirely dependent on the solubility of H2S  in

water.  In all cases, where concentrations above 100 ppm H2S existed, H2S

did evolve from the solution.
                                      42

-------
          50
          40
          10
•
                                    j
             0         0.1          0.2
                     Absorbance

Figure 12- Calibration Curve for t^S Using Methylene Blue-Spec 20
                               43

-------
   500
   400
   300
a.
a
10

CN



U
   200
   100
                0.1
                                    SPECTROPHOTOMETER - 20

                                    B&L. FILTER  670 A
                           _L
0.2        0.3

 ABSORBANCE
0.4
0.5
Figure  13 -  Calibration Curve for lUS  Using Methylene Blue-Spec 20
                                    44

-------
   500 _
   400
   300
 I
CN
X.
(J
   200
    100
                 Jl
                         SPECTROPHOTOMETER - 20
                         B& L. FILTER   670 A
I
           I
                0.1        0.2       0.3

                       ABSORBANCE
                    0.4
  Figure 14- Calibration Curve for t^S  Using Methylene Blue-Spec 20
                                 45

-------
          Based upon the work at the refinery and at MRI in trying to





establish calibration curves for the methylene blue-spectrophotometric





method for analyses of H2S, the conclusion is that the method is not appli-





cable for concentrations above 100 ppm of H2S and is of questionable value




above 50 ppm H2S.







                  V.  SAMPLING AND ANALYTICAL PROCEDURES







          Figure 15 is a schematic of the equipment used to obtain the gas




samples.  A pressure reducing valve with all Teflon contact parts was in-




stalled in the outlet of a bleed valve on the low pressure line from the




amines scrubbers.  A 1/8-inch Teflon tube about 40 feet long was attached




to the outlet of the pressure reducing valve.  The Teflon tubing was jacketed




with a 3/8-inch copper tubing to provide steam tracing and also prevent





damage to the line.  A Teflon tee was put into the sample line 40 feet from





the pressure regulator.  One branch of the tee went to the gas chromatograph




sample valve.  The other end of the tee was connected to the sampling train.




The sampling train consisted of five impingers in series, 15 ml of cadmium




hydroxide solution in the first two, two dry impingers and a silica gel




impinger.  The impingers were submerged in an ice water bath to be certain





that all of the moisture in the line was condensed in the dry impingers and





the silica gel impinger before the gas was discharged to the dry gas meter.
                                   46

-------
Refinery
Gas line
From Scrubber
Pressure
Reducing
Valve
                         l/8"Teflon Tubing
                         Copper Jacket
                     Pressure Gauge
                  Therometer  -_ ^_^^
                         "99
                        2 - Cd (OH)2 Impinger

                        Dry Impinger

                       Silica Gel

                        (      r^r^TMTKT^
                             Drygas
                             Meter
                                                                    To G.C.  Sample Valve
                                                                      Ice Water Bath
                                    Impinger
                                    Train
                             Figure 15 - Schematic of Sampling Equipment

-------
          The EPA suggested sampling procedure is contained in Appendix A





for methylene blue and Appendix B for iodometric titration.  These sampling





procedures were modified by the addition of three impingers,  two dry and one




filled with silica gel following the two impingers containing 15 ml each of





cadmium hydroxide solution.





          When the absorbing solution is made up, the cadmium hydroxide





solubility is so low that it precipitates out and forms a white crystalline




precipitate, this should be mentioned in the procedure for preparation of




the absorbing solution.  It also should be stressed, that before the absorb-





ing solution is poured into the impingers the bottle of absorbing solution





must be thoroughly shaken to make sure that the precipitated hydroxide does




actually get into the impingers.




          The procedure specifies a sampling rate of Z-rl/2 liters/minute




for 10 minutes, for a total of 25 liters of gas sampled.   However, the first




day, it was found that 25 liters of gas, at the concentration of H2S in




this refinery stream, was exhausting the 30 ml of cadmium hydroxide before




the end of the sampling period.  At the concentration of l^S (200 to 500 ppm)




in the refinery gas stream, an adequate sample volume was 10 to 15 liters




of gas.  Using this sample volume, the cadmium hydroxide was not exhausted




during sampling.





          At the start the procedure as presented in Appendix A, for the




Recovery of the Collected Sample, was followed.  But a problem arose in





transferring the contents of the impingers quantitatively to a plastic




storage bottle.  The washing of each impinger and the connecting




                                    48

-------
glassware with the distilled water did not remove all of the precipitated





cadmium sulfide present in the impingers and the connecting glassware.





          Attempts were made to rinse the glassware with 10% by weight  HC1,





but loss of H2S when the CdS dissolved was observed.





          According to the proposed method in Appendix B, the CdS precipitate





was filtered through a Whatman No. 40 filter paper, but this did not work.





The filter paper did not retain the cadmium sulfide.  Filtering was then




tried through glass crucibles with the porous plates (filter crucible)  and




rinsing the sample container and the precipitate with 10% ammonium hydroxide




solution.  This was not successful either, due to the inability to recover





all the CdS from the glassware and the inability of the crucible to retain




the precipitate.




          Additionally, the retained filter cake could not be dissolved




and recovered because of the loss of H^S.




          Because of these problems, the intermediate filtering step was




discarded and sample recovery was accomplished according to the procedures





given in the suggested method outlined in Section  III-B.




          The method using the iodometric thiosulfate titration is a usable.




method as long as the analysis is performed within 1 hour of the sampling





time.  Transportation of the samples back to a central lab for analysis is




impractical unless it is transported back in the same container that the





sample was obtained in.  In other words, it would have to be transported





back in the midget impingers and analyzed in the midget impingers at the Iab0








                                    49

-------
The preferred method is to take the sample and dissolve the cadmium hydroxide





in the impingers and then transfer qualitatively the dissolved materials to





the iodine flask.  Once the cadmium sulfide has been dissolved in the acidi-





fied iodine solution, the analysis should be completed within 30-45 minutes





before the loss of iodine begins to be troublesome.





          Figure 16, is a schematic of the analytical setup for the gas




chromatographic flame photometric analysis of H2S.




          The sampling for the GC consisted merely of injecting gas from the




refinery gas stream into the sample valve and from the sample valve directly




to the gas chromatograph column into the flame photometer and detector.





The GC is a Perkin Elmer Mark II* with a  stainless steel gas sampling valve,




Melpar Model FTD100* flame photometric detector, a Microtech Model 8176*




solid state electrometer, and a Leeds and Northrup Speedomax XL* recorder.




Nitrogen was used as the carrier gas and a mixture of hydrogen, oxygen,




and air was burned in the flame photometer.





          Due to the mutual solubility and reaction of methylethylamine




and H2S problems were experienced in the operations of the gas chromatograph.




The Teflon column became saturated with methylethylamine and would not elute




H2S to the detector.  In the future for GC work, it would be wise to include





a column to scrub out the methylethylamine ahead of the 36-foot Teflon




column used for hydrogen sulfide resolution.
*  Mention of a company name does not indicate endorsement, by EPA.






                                   50

-------
   STANDARD
H2S 46ppm IN N2
            Sample
            .Line
                                               rtf
                                N2
         AIR
       25 cc/mi
                          -33-
                                                               15 cc/mln
mm
                              Flame Photometer and Detector
               Sample  Gas Chromatograph
               Valve       Column
Electrometer
                                                                              70 cc/min
                                                                                            Power Regulator
           Flowmeter
           Assembly
                                                                                                             J110 Volt
Recorder
                                          Figure 16 - Gas Chromatograph Analytical Train

-------
          Figures 17 through 26 are photographs taken at the refinery





during the sampling program.  Figure 17 shows the sampling train with the





impingers in the ice water bath, the Teflon sample line and the dry gas




meter for recording the temperature and amount of gas sample.





          Figure 18 shows the Teflon sampling line to the gas sampling





valve on the chromatograph as well as the general layout of the chromato-




graphic analytical equipment.




          Figure 19 is a side view of the gas chromatograph showing the




set up, and Figure 20 is the view of the gas chromatograph showing the




cylinder of the l^S standard gas, the sample valve, the gas chromatograph,




the flame photometer, and detector, the electrometer, the flowmeters and




the recorder integrator and the power supply regulator.





          Figures 21 through 25 show the sequence of analysis using the




iodometric titration.  Figure 21 shows the impingers with the yellow cad-




mium sulfide precipitate, Figure 22 shows the flask with the dissolved





sulfide in the acidified iodine before titration.  Figure 23 shows the




flask during titration and before the addition of the starch indicator.




Figure 24 shows the flask after addition of the starch indicator, and




Figure 25 shows the flask at the end point of the iodometric titration.





Figure 26 shows the Bausch and Lomb Spectral Photometer Model No. 20* with




a blank and the H2S methylene blue solutions ready for analysis.
*  Mention of a company's name does not indicate endorsement by EPA.
                                   52

-------
Figure 17 - Impingers  and Dry  Gas Meter
     for Hydrogen Sulfide Sampling
Figure 18 - Teflon Sample Line to Gas
         Sampling Valve on GC
                    Figure 19 - Side View of Gas Chromatograph
                                      53

-------
Figure 20 - Gas Chromatograph Setup
  Showing - H2S Standard,  Sample Valve -
  GC-Flame Photometer-Electrometer-Flow
  Meters-Recorder Integrator
Figure 21 - Impingers with CdS
                         Figure 22  - Flask with CdS  - Acidified
                                 Iodine  Before Titration
                                        54

-------

Figure 23 - Flask During Titration Before
       Addition of Starch Indicator
 Figure  24  - Flask After Addition of
            Starch Indicator
 Figure  25  -  Flask  Showing End Point of
         lodometric Titration
Figure 26 - B and L Spec 20 with Blank
  and HoS-Methylene Blue Solutions
                                       55

-------
          VI.  PROCESS OPERATING CONDITIONS
This section to be furnished by EPA.
                          56

-------
                               APPENDIX A
                   DETERMINATION OF H2S IN STACK GASES
          The following represents the draft method of sampling for hydrogen




sulfide and analyzing by the methylene blue-spectrophotometric method.  This




method was furnished by EPA.  Observations on this method and the revised




method are tabulated in Section IV.
                                    57

-------
                   DETERMINATION OF H2S IN STACK GASES







1.  Principle and Applicability





     1.1  Principle




          H2S is collected in a series of midget impingers and reacted with





alkaline Cd(CH)2 to form CdS.  The CdS is filtered and rinsed with cold water.




It is then dissolved in HC1 and reacted with N,N dimethyl - p-phenylenediamine




in sulfuric acid to produce methylene blue, which is determined spectrophoto-




metrically at 670 mm.




     1.2  Applicability





          This method should be applied only when specified by the test




procedures for determining compliance with the New Source Performance Standards.







2.  Range and Sensitivity




          This method is reported as usuable down to 5 ppb, and to beyond




500 ppm.







3.  Interferences




          SOo in concentrations 100 times greater than HoS is reported to





cause some interference.







4.  Apparatus




     4.1  The apparatus used shall be the same as that used in Method 4,




F.R. 36, 24887, 23 December 1971.




     4.2  Spectrophotometer shall be capable of operation at 670 nm.
                                     58

-------
5.  Reagents

     5.1  Sampling

          5.1.1  Absorbing solution - Mix 4.3 g cadmium sulfate (30dSO^

8H20) and 0.3 g of NaOH in 1 liter of distilled water.  Mix well before

using.

     5.2  Analysis

          5.2.1  Amine-acid stock solution - Add 50 ml of concentrated

sulfuric acid to 30 ml of water and cool.  CAUTION:  Add H2S04 slowly or

spattering may result.  Add 2 g of N,N-dimethyl - p-phenylenediamine.  Stir

until dissolved.

          5.2.3  Ferric chloride solution - Dissolve 100 g of ferric chloride

hexahydrate (Fe2Cl3 - 6H20) in water and dilute to 100 ml.

          5.2.4  4 N HC1 - Dilute 340 ml of concentrated HC1 to 1 liter.

                        *
6.  Procedure

     6.1  Sampling

          6.1.1  Assemble the sampling apparatus as shown in Figure 4-1,

F.R. 36, 24887, connecting two midget impingers in series.  Transfer 15 ml

of absorbed solution to each impinger.

          6.1.2  Sample at a rate proportional to the stack velocity until

a 25-liter sample has been obtained.

          6.1.3  Transfer the contents of the impingers quantatively to a

plastic bottle, washing each impinger several times with distilled HoO.
                                    59

-------
     6. 2  Analysis




          6.2.1  Filter the contents of the sample container 'through each




of two sintered glass crucibles.  Rinse several times with 2 ml portions of




cold water.  Discard the effluent.  Dissolve the CdS in 4 H HC1 and immediately




add 3 ml of amine test solution and one drop of ferric chloride solution.




Agitate after each addition.  Transfer to 50 ml volumetric flasks, make up




to volume, and allow to stand for 30 minutes.  Run a blank in the same




manner using unaspirated absorbing solution.  Determine the absorbance of




the sample at 670 nm vs the blank.  Determine the l^S concentration from a




previously prepared standard curve.






7.  Standardization




     7.1  Using analyzed reagent grade CdS, prepare a working calibration




curve starting at Section 6.2.1  of the procedure.  The calibration curve




should have a minimum of five points.






8.  Calculations




             CH2S = Ccds x VSOLN x 135756 X °-0154 gr/mg X Vj





                CT, g = Concentration of H^S in gas analyzed




                Cr,c = Concentration of CdS from calibration curve in mg/ml
                 Lido



                Vg   = Volume of gas sample in SCF




                     = Fi-nal dilution volume of solution read in spectrophotometer
                     = Ratio of mol. wt H0S/mol wt CdS.
              144.46
                                    60

-------
9.  Bibliography
Jacobs, M. B., M. M. Braverman, and  S. Hochheiser.   "Ultramicrodetermina tion
of Sulfide in Air," Anal. Chem..  29,  1349  (1957).
                                     61

-------
                               APPENDIX B







                DETERMINATION OF  H2S IN REFINERY FUEL  GASES








          The following represents the draft method of sampling for hydrogen




sulfide and analyzing by the iodometric titration method.  This method was




furnished by EPA.  Observations on the method and the revised method are




tabulated in Section IV.
                                     62

-------
               DETERMINATION OF H2S IN REFINERY FUEL GASES
1.  Principle and Applicability





     1.1  Principle




          H2S is collected in a series of midget impingers and reacted with





alkaline Cd(OH)2 to form CdS.  Then CdS is filtered and rinsed with 10% NH.OH.





It is then dissolved in HC1 and absorbed in a known volume of iodine solution.




The excess iodine is titrated with thiosulfate solution, with starch as the




indicator.




     1.2  Applicability





          This method should be applied only when specified by the test




procedures for determining compliance with the New Source Performance




Standards.







2.  Range and Sensitivity







3.  Interferences







4.  Apparatus




     4.1  Sampling





          The apparatus used shall be the same as that used in Method 4,




F.R., 36, 24887, 23 December 1971.




     4.2  Analysis





          Glass stoppered 500 ml iodine number flask for reaction vessel.
                                    63

-------
5.  Reagents





     5.1  Sampling





          5.1.1  Absorbing solution.




          Mix 4.3 g cadmium sulfate hydrate 3 CdS04'8H20 and 0.3 g of NaOH





in 1 liter of distilled H-O.  Mix well.





     5.2  Analysis




          5.2.1  10% Ammonium hydroxide solution




          Add 380 ml of concentrated ammonium hydroxide (s.g. 0.90) to




620 ml of distilled H20.




          5.2.2  10% Hydrochloric acid solution.




          Mix 230 ml of concentrated HC1 (s.g. 1.19) and 770 ml of distilled




H20.




          5.2.3  Iodine solution, O.lN.





          Dissolve 25 g of potassium iodide (KI) in 30 ml of distilled H20




in a 1 liter graduated cylinder.  Weigh  12.7 g of resublimed iodine (I2)




into a weighing bottle and add to the potassium iodide solution.  Shake the





mixture until the iodine is completely dissolved.  Slowly dilute the solu-




tion to 1 liter with distilled 1^0, with swirling.  Filter the solution, if




cloudy, and store in a brown glass stoppered bottle.




          5.2.4  Iodine solution, 0.01 N.




          Dilute to 1 liter with distilled water 100 ml * 0.01 ml of the




0.1 N iodine solution in a volumetric flask.
                                     64

-------
          5.2.4  Iodine solution, 0.01 N.(Continued)





          Standardize daily as follows:  Pipette 100 ml of the 0.01 N iodine





solution into a 500-ml conical flask.  Titrate with the standard 0.01 N thio-





sulfate solution until the solution is a light yellow.  Add a few drops of





the starch solution and continue titrating until the blue color just





disappears.





          5.2.5  Sodium thiosulfate solution.




          Sodium thiosulfate solution, standard 0.1 N.  For each liter of




solution, dissolve 24.8 g of sodium thiosulfate (^28203'51^0) in distilled




water and add 0.01 g of anhydrous sodium carbonate (Na£ CO-j) and 0.4 ml of




chloroform (CHC13) to stabilize.  Mix thoroughly by shaking or by aerating




with nitrogen for approximately 15 minutes, and store in a glass-stoppered





glass bottle.




          Standardize frequently as follows:  Weigh into a 500-ml volumetric





flask about 2 g of potassium dichromate (lOjC^Oy) weighed to the nearest




milligram and dilute to the 500-ml mark.  Use dichromate which has been




crystallized from distilled water and oven-dried at 360°F to 390°F.  Dissolve





approximately 3 g of potassium iodide (KI)  in 50 ml of distilled water in a




glass-stoppered, 500-ml conical flask, then add 5 ml of 20% hydrochloric




acid solution.  Pipette 50 ml of the dichromate solution into this mixture.




Gently swirl the solution once and allow it to stand in the dark for 5 minutes.




Dilute the solution with 100 ml to 200 ml of distilled water, washing down





the sides of the flask with part of the water.  Swirl the solution slowly





and titrate with the thiosulfate solution until the solution is light yellow.




                                     65

-------
Add 4 ml of starch solution and continue with a slow titration with the




thiosulfate until the bright blue color has disappeared and only the pale




green color of the chromic ion remains.




          5.2.6  Sodium thiosulfate solution.




          Sodium thiosulfate solution, standard 0.01 N.  Dilute 100 * 0.01 ml




of the standard 0.1 N thiosulfate solution in a volumetric flask to 1 liter




with distilled water.




          5.2.7  Starch indicator solution.




          Suspend 10 g of soluble starch in 100 ml of distilled water and




add 15 g of potassium hydroxide pellets.  Stir until dissolved, dilute with




900 ml of distilled water, and let stand 1 hour.  Neutralize the alkali with




concentrated hydrochloric acid, using an indicator paper similar to "alkacid"




test ribbon, then add 2 ml of glacial acetic acid as a preservative.




          Test for decomposition, by titrating 4 ml of starch solution in




200 ml of distilled water, with the 0.01 N iodine solution.  If more than




four drops of the 0.01 N iodine solution are required, make up a fresh




starch solution.







6.  Procedure




     6.1  Sampling




          6.1.1  Assemble the sampling train as shown in Figure 4-1, F.R. 36




24887 connecting two midget impingers in series.  Place 15 ml of the collec-




tion solution in each impinger.
                                   66

-------
          6.1.2  Purge the connecting line between the fuel gas sampling




valve and the first impinger.  Connect the sample line to the train.  Read





and record the initial reading on the dry gas meter.




          6.1.3  Open the flow control valve and adjust the sampling rate





to 2.5 liters/min.  Read and record the meter temperature.




          6.1.4  Continue sampling for 10 minutes.  At the end of this time,





close the flow control valve and read the final meter volume and temperature.




          6.1.5  Disconnect the impinger train from the sampling line and




cap the open ends.  Remove to the sample clean-up area.





     6.2  Sample Recovery




          6.2.1  Transfer the contents of the impingers quantitatively to a




plastic storage bottle.  Wash each impinger and the connecting glassware





several times with distilled water, and add these washings to the storage




container.  Cap and seal the storage container.




     6.3  Analysis




          6.3.1  Filter the contents of the sample container through each of




two retentive filter papers,  (eq. Whatman 40).  Rinse the sample container




and the precipitate twice with 10% ammonium hydroxide solution.





          6.3.2  Place the filter papers with the precipitate in the iodine




number flask.  Add a measured excess of Standard 0.01 N iodine solution and




immediately add 50 ml of 1070 HC1 solution.  Stopper and shake.  (Enough




standard iodine solution should be added to give a back titration equivalent





to approximately 50% of the added iodine  solution.)
                                    67

-------
          6.3.3  Titrate the solution in the flask with 0.1 N sodium thio-
sulfate solution until the solution is light yellow.  Add 4 ml of the starch
solution and continue titrating until the blue color just disappears.
          6.3.4  Run a blank determination beginning at 6.3.1 to adjust the
value of iodine consumption.

7.  Calculation
     7.1  Gas volume sampled at standard conditions (70°F, 29.92 in Hg)

                       v   = 17.7 	5n	 V   ft3
                                  (Tm + 460)  m

                        Pm = meter pressure "Hg abs
                        Tm = meter temperature, °F
                                             •3
                        V_ = meter volume  ft
                             0.2618(AN, - BN,)
                      CH2S = 	=—-1	 gr/SCf
                                   ms
                      where
                         A = volume of standard iodine used, ml
                        N- = normality of the standard iodine solution
                         B = volume of standard sodium thiosulfate used, ml
                        N£ = normality of the standard thiosulfate solution
                       Vms = volume of gas sampled at standard conditions

                  0.2618 =  (34 g/mole H2S) (0.0154 gr/mg) (1000 mg/g)
                               (1000) (2 H2S equivalent)
                                    68

-------
                     APPENDIX C
                  FIELD DATA SHEET
The Field Data Sheets are contained in this Appendix.
                         69

-------
                         GAS SAMPLING FIELD DATA
Material Sampled for_
Date
Plant
Ambient Temp




Run No.
      /(£>
/i
Power Stat Setting
Filter Used:  Yes
Operator
                      No
                                   Location
                               F   Stack Temperature




                                   Stack Dimensions
                    Irapingers with
                    Ini)ingers with
                    Total number of Impingers




                    Sample Bottle No.
                    Impinger Bucket No.
                    Meter Box No.    U'*2.
                                                  ml of
                                                                         °F
                                                              .£    y/4-L \/
                                           \/    ml of  /
                                           -
                                        70

-------
  Clock
  Time
  Meter (ft3)
 Flow Meter
Setting (cfh)
Meter Temperature
     in
             o.ooo
 /ooo
                           - A
              /   p
              '/<*> ra*.
                                                  77
            0.321
                                                  71
 1/3- 1
                      0
  nsv
0.3*7
            n.ooo
\
             O.&&0
  134-7
 0.323
 2.0
  1411
             0 .BOO
                             2.0
                             3.0

-------
                         GAS SAMPLING FIELD DATA
Material Sampled for	




Date   <£ $  J ifflf  72-
Vs<-  —
Plant  UW/DAI Vs<- —*vL/&T_£ c Location
Ambient Temp.     "7J 	°F   Stack Temperature_




Run No.,^?'^  'T&S1'  F}/)f	   Stack Dimensions	




Power Stat Setting	




Filter Used:  Yes	No




Operator
                    Impingers with _    _ ml of
                    Impingers with _ ml of     / c
                    Total number of Impingers




                    Sample Bottle No.     —
                    Impinger Bucket No.




                    Meter Box No.

-------
Clock
Time
              Meter (ft3)
 Flow Meter
Setting (cfh)
Meter Temperature
    in
 //ft
                O.ooo
 lot
                                                  73*
           0. QOO
                    n
           O.ooo
          O . & b O
   /7
              O.f/7
                            •  o
                                                   7
                          3.o
                                                   7/
          0.313
          O. 6x30
1
          Q.232-
                                                 72*
                          2.0
           n.ooo
                          3.0
/**>/
A//
                              ^L
                       23.
          £.
                              ^_0
          O.M'
                                73

-------
7-2-
'e^f
f/*
&<
M// '
j
Clock
Time
/Ct2f
' U&
J7/0
n^o


















•

Meter (ft3)
O BOO
O 3 2^=*
0.00O
0. 3^P-




















Flow Meter
Setting (cfh)
3-.o
3.o
a~*
<3.
-------
                         GAS SAMPLING FIELD DATA
Material Sampled for_
Date
Plant
[}
to i /
             Location
Ambient Temp .    G? f?   f~    °F   Stack Temperature
Run No.
             Stack Dimensions
Power Stat Setting
Filter Used:  Yes
Operator_
No
         I
                    Impingers with
                                                               J-
                    Impingers with
                    Total number of Impingers_




                    Sample Bottle No.	
                    Impinger Bucket No.	




                    Meter Box No.	Cs
                            ml of
                            ml of
                       (/     '
                                         75

-------
                                 92-
 Clock
 Time
              Meter (ft3)
           0.000
           0.33^
 Flow Meter
Setting (cfh)
                              Jo
                              _£0_
Meter Temperature
    in
oQfo
          0.3V/
           0.000
          fi.OOO
                              -2. ?
/of?
 1//7
         O.3S3
          0.31^
*J_jtW
         o.ooo
         o,
                             s.i
                        3.0
//a./
          0.
                            /,
 \33o
Mo*
         D.&ev
             0.301

-------
     Clock
     T±me
Meter (ft3)
 Flow Meter
Setting  (cfh)
Meter Temperature
     in
                   .00°
                                       73 '
                                                       fa
     /fro
o.
                 0.000

                                                oo
                0.3
                 2.3
T&Vf
                 I.I
               0*3 80
                1L
                O.QQO
                                 1.4-
     ll /
               0.302.
                                 2.1
                                 3.1
                                           77

-------
                         GAS SAMPLING FIELD DATA
Material Sampled for




Date   30
                         JL
      Ifr/y
Plant L/AJ/osJ  ([//(-- ^oL^^'f   Location	




Ambient Temp.	X"«J	°F   Stack Temperature
Run No. _ . _   Stack Dimensions




Power Stat Setting    —  ~" _




Filter Used:  Yes     No  fX^




Operator
                    Impingers with     / S"~ _ ml of
                                                        CJ **H )-,
        / _ Impingers with
                    Total number of Impingers_




                    Sample Bottle No.	—
                    Impinger Bucket No.




                    Meter Box No.	Cf "~
                                         78

-------
                             •30
•"£*/
     Clock
     Time
 Meter (ft3)
                              Flow Meter
                                   (cfh)
Meter Temperature
    in
               0-000
      owe
030+
                               A3.
      09.29
Q.OOO
                              2.3-
    79
              o.o&o
                               1.1
              O
                              f.e
                               3.S
                               2.3.
                               3.7
f?
^'
              0.3 t,l
                1?
     Jtt?-
              O.Zf?
              4). OOo
                  ?
      130(0
     ^_
              ^
               fl-Q^o
              ^i£
               V.ZJt,
                                    79

-------
                                 72-
 Clock
 Time
              Meter (ft3)
 Flow Meter
Setting (cfh)
Meter Temperature
    in
w
  4)2-
               0.309
                                            'f
          0.000
 14-31
               a
  /.f
            . wo
          0.349
                              3.1
                         3.1
          0.
                             2.
          Q.QOO
                                                f
          o.
          O'.ooo
         O.ooo
                             I.-J
         0.3o4
                                            xrf
                          5.
                                            SZ'f
                                       tl'f
/Iff
                            sn
                  f
        0,
                            sn
               t&f
        0.273
                               80

-------
                               APPENDIX D


                           LABORATORY REPORT
                           (Titration Data)


          This section presents the iodometric titration and the methylene

blue-spectrophotometric field data sheets.
                                    81

-------
                                                                               //
                                                                                       -?
00

-------
/
     n,l
                                  /
    "r .
                 &< O/
2-3,0
           83

-------
                             /I
O
             84

-------
85

-------
&^
                                  7:3
                 86

-------
                               72^
 IAJ$
^o s ^ //
    2
        /?'•

                           IhtO
                           f /:3 5
              87

-------
     -*
J^^y?A

              sr.B
Scv^/e^JL/J



     ^
               3O.T
             4-7,5
                             fl?
                    88

-------
1S '3
                    7 d : S'O
                    •//;
                          : 3
        89

-------
-*5**°fl&&$£
            /
-------
91

-------
7
              92

-------
 (C:
93

-------
94

-------
i«j£ "<3+~^t, ~&^s <£&?//
          ^
               95

-------
    4 2.
  -30(0  -
                                              / :
I «/ 6" **> -  / jT^ 3
      ,   /
                                96

-------
/ <= O
                      y
               &./&>  .
            97

-------
                               APPENDIX E







                               FIELD LOG







          Table E-l presents the actual time during which sampling was




conducted.
                                    98

-------
                               TABLE E-l

                              SAMPLING LOG
                                                               Time
Location

Amine scrubber effluent
low pressure line

Amine scrubber effluent
low pressure line

Amine scrubber effluent
low pressure line
Pollutant
H2S
CH3-SH
H2S
CH3'SH
H2S
Date
6/27/72
6/28/72
6/29/72
Began
10:30
11:25
9:06
Ended
15:35
17:20
16:42
CH3-SH
Amine scrubber effluent
low pressure line
  H2S
 CH3-SH
6/30/72
8:53   18:00
                                  99

-------
                          APPENDIX F
               PROJECT PARTICIPANTS AND TITLES
    Name





Paul Constant




Evan P. Shea




Gary Kelso




Fred Bergman





Doug Weatherman





Mike Serrone
       Title




Program Manager




Project Chief




Testing Engineer (gas)




Analytical Chemist (GC)




Analytical Chemist




Analytical Chemist
                                100

-------