xvEPA
          United States
          Environmental Protection
          Agency
          Office of Air Quality
          Planning and Standards
          Research Triangle Park NC 27711
EMB Report 78-OCM-3
March 1979
          Air
Benzene Organic
Chemical Manufacturing

Emission Test Report:
Ethylbenzene/Styrene
Cos-Mar
Carville, Louisiana

-------
         SOURCE TEST AT COS-MAR'S

        ETHYLBENZENE/STYRENE PLANT

            CARVILLE, LOUISIANA
          Contract No. 68-02-2812
              Work Assignment 11
   EPA Technical  Manager:  Winton Kelly
               Prepared for:

    U.S.  ENVIRONMENTAL PROTECTION AGENCY
.Emission  Standards  and Engineering Division
        Emission   Measurements Branch
Research  Triangle Park, North Carolina  27711
                   TRW
         ENVIRONMENTAL ENGINEERING DIVISION

-------
                         TABLE OF CONTENTS
Section                                                             Page
1.0  INTRODUCTION 	         1

2.0  SUMMARY OF RESULTS	        -2

3.0  PROCESS DESCRIPTION  	        22

4.0  LOCATION OF SAMPLING POINTS  	        23

5.0  SAMPLING AND ANALYSIS PROCEDURE  	        29

APPENDICES
A.  COMPLETE RESULTS AND CALCULATIONS 	        36

B.  GAS CHROMATOGRAPH RESULTS	        41

C.  LABORATORY RESULTS	•	        63

D.  SAMPLING PROCEDURES 	       115

E.  FIELD DATA SHEETS.	'.	       121

F.  TEST LOG	       154

G.  PROJECT PARTICIPANTS  	       157

-------
                             LIST OF FIGURES

Number                                                       Page
2.1  Benzene Drying Column Vent Equipment 	    6
2.2  Modified Moisture Train  	    7
2.3  Caustic Scrubber Equipment 	    8
2.4  Catalyst Mix Tank Equipment	    9
2.5  Benzene/Toluene Column Vacuum Equipment  	   10
4.1  Benzene Drying Column Vent Equipment 	   25
4.2  Caustic Scrubber Equipment 	   26
4.3  Catalyst Mix Tank Equipment	27
4.4  Benzene/Toluene Column Vacuum Equipment  	   28

-------
LIST OF TABLES
Numbei
2.1
2.2
2.3
2.4
2.5
2.6
2.7
2.8
2.9
2.10
2.11
P
Results of Gas Sampling Analysis - Test Point #1 ....
Results of Gas Sampling Analysis - Test Point #2 ....
Summary of Results Species Anaylsis - Organic
Condensate at Points 1 and 2 	
Results of Gas Sampling Analysis - Test Point #4 ....
Results of Gas Sampling Analysis - Test Point #5 ....
Summary of Results Caustic Scrubber Liquid Analysis
(Point #6) 	
Results of Gas Sampling Analysis - Test Point #7 ....
Results of Gas Sampling Analysis - Test Point #8 ....
Results of Gas Sampling Analysis - Test Point #9 ....
Results of Gas Sampling Analysis - Test Point #10. . . .
Gas Flow Calculations for Locations 5, 7, and 10 ....
'age
11
12
13
14
15
16
17
18
19
20
21

-------
                          1.0  INTRODUCTION

      Under the Clean Air Act, The U.S.  Environmental  Protection Agency
(EPA) is required to establish National  Emission Standards for Hazardous
Pollutants for emissions that have been  found to cause adverse health effects.
Benzene has been listed as a hazardous pollutant and studies have been ini-
tiated to develop background information.   The test program at this facility
was conducted to collect emission data from ethylbenzene-styrene production
for these studies.
      Testing was conducted at the Cos-Mar, Carville,  Louisiana No. 2 Ethyl-
benzene-Styrene plant during June 19-30  and July 10-14, 1978, By TRW Environ-
mental Engineering Division personnel  under contract to EPA.  Testing was
coordinated and observed by a representative of Monsanto Research Corporation,
also under contract to EPA.
      The purpose of testing was to obtain data before and after control
devices (if present) for total organics  and specifically, benzene from the
following systems:
      - Benzene Drying Column Vent Equipment
      - Alkylate Degasser Vent Equipment
      - Catalyst Mix Tank Equipment
      - Benzene/Toluene Column Vent Equipment
      The number of locations sampled, a system description, summary of
specific test methods, sample point locations, and a results presentation are
given separately in Section 2 of this  report for each  of the above systems.
General test methods are presented in  Section 5.  Detailed sample location
descriptions are included in Section 4.   Complete data summaries, sample
calculations, and field data are included in appendices A through G.

-------
                          2.0  SUMMARY OF RESULTS

 2.1  BENZENE DRYING COLUMN EQUIPMENT
    The benzene drying column vent system is illustrated in Figure 2.1.  A
 condenser is used to further remove organics from the vent streams from the
 column reflux condenser and reflux decanter.  The exit stream from this
 condenser is routed to the plant flare system.  Sampling was conducted before  ,
 and after this condenser at Points 1 and 2 as indicated in Figure 2.1.
    Gas samples were collected simultaneously into flexible plastic bags
 (Tedlar) at each location.  The system pressure (80 psig) was used to fill
 the bags.  Samples were extracted through existing side-tap valves on the
 piping.  During initial sampling attempts, condensate was observed in the
 sample bags.  In an attempt to avoid this problem, a condenser-knockout system
 was added prior to the sample bag.  This system consisted of the equipment
 normally used for moisture determination operated in an ice bath (see Figure
 2.2).  This modification prevented the appearance of condensate in the sample
 bags.
    A two-phase liquid was observed in the condenser train after sampling.
 The lower phase was water and  was used for calculation of the stream
 moisture content.  The top phase represented condensed organics and, possibly,
 liquid entrainment from the stream sampled.  It was observed that liquids were
 entrained in the flowing streams at both points.  It is not possible to predict
 how much of the collected condensate is attributable  to trapped entrained liquids
 and how much was due to condensation.
    This liquid was recovered and analyzed.  However, no attempt was made to
 combine these results with the vapor sample results to obtain total stream
 organic content.
    Analyses were conducted on the bag samples to determine nonaromatics  as
 GI - GS (species) and aromatic compounds, specifically, benzene,  toluene,
xylene, ethyl benzene, and styrene.
    An orifice flow meter was located near'Point 2.  This meter was to be
 used for flow data, but the presence of entrained liquids in the stream
 resulted in erratic differential pressure levels and prevented-the measurement
 of flow rates.

-------
      The results of gas sampling are presented in Tables 2.1  and 2.2.   As
discussed previously, these data represent the vapor phase compounds present
after a condenser (i.e., impingers) and are potentially less than the concen-
trations existing in the streams.  This would be due to the difference
between the source and the final condenser train temperature.
      The vapor sampling results are presented for C-j - 65 nonaromatic
compounds and the indicated aromatics on the basis of benzene equivalents
and as the specified compound.  Two results are presented for C4 because two
defined peaks were present with retention times near that of butane.  Addi-
tional studies were not performed to identify these compounds  specifically.
      The results presented for total hydrocarbons by total HC analyzer are
not representative of true concentrations.  Because of the high organic
concentration, the detector of the instrument was saturated.  In other words,
the concentration was beyond the analyzer range, and all  results represent
the upper limit measurable.  A dilution apparatus was not available; therefore,
the sample concentration could not be reduced to a level  where valid results
could be obtained on a total HC analyzer.  The total hydrocarbons by summation
of the individual species results should represent the total organics concen-
tration in each stream.
      On Run 2-2, the results for CB and C$ are significatly different than
the results for the other two runs.  Analysis records were rechecked to
comfirm the results.  No reason is known for the apparently different results.
The total organics in the condensate are presented in Table 2.3.

2.2 ALKYLATE DEGASSER VENT EQUIPMENT
      The alkylate degasser vent equipment is illustrated in Figure 2.3.
The vent stream from the degasser is first scrubbed with a polyethylbenzene
solution and then is routed to a caustic scrubber along with a vent from
other equipment.  During the test period, there was no flow from the other
equipment (Test Point 3); therefore, no measurements were performed at that
location.  After caustic scrubbing, the vent stream is routed  to the plant
flare system.  The caustic solution charge to the scrubber vessel is on a
batchwise basis.  During sampling, no fresh caustic was added.
      Gas samples were collected from the streams before and after the caustic
scrubber.  Grab samples of the caustic liquid were also collected during gas
stream testing.  The grab samples of caustic liquid were analyzed for

-------
Na+ and Cl" and the results are listed in Table 2.6.   Integrated samples were
collected in flexible bags for organic compound analysis and inert gas
analysis.  In addition, samples were collected from the gas stream using a
condenser train for moisture content determination.  Finally, gas samples were
extracted using the procedures given in Section 5 for HC1 determination.  The
flow rate at the scrubber outlet was measured using a vane anemometer attached
to the outlet of the vent to atmosphere.  (Note:  For sampling purposes, the
scrubber outlet stream was not routed to the flare system.  Instead, it was
bypassed through an existing vent to atmosphere.)  The inert gas analysis can
be used to estimate the inlet flow rate using a nitrogen balance.
     The results of gas stream analysis are summarized in Tables 2.4 and 2.5.
The results of caustic liquid analysis are presented in Table 2.6.
     At Point 4, there are significant differences in the results for 04, C$,
benzene,and ethyl benzene between the two runs.  Also, at Point 5, there are
significant differences in (4, C^s.and benzene.  The analysis records were
rechecked to verify the results.  One reason for the differences may be due to
the fact that the two runs were separated by about two weeks.
2.3  CATALYST MIX TANK EQUIPMENT
     The catalyst mix tank vent equipment is illustrated in Figure 2.4.  The
equipment included is the catalyst storage vessel and the mix tank where
catalyst is dissolved in a polyethlybenzene solution.  The vent gasses are
essentially nitrogen purge streams that are used to inert the system.  The
vent stream is scrubbed with fresh polyethylbenzene solution prior to discharge
to the atmosphere.
     Samples were collected for organic species analysis and to determine HC1
content of the vent stream.  The vent flow rate was determined using a vane
anemometer.
     Sampling was conducted during catalyst mixing only, which is a batch
process activated on an as-needed basis.
     The results of testing are summarized in Table 2.7.
2.4  BENZENE-TOLUENE COLUMN VACUUM EQUIPMENT
     The vacuum equipment serving the benzene-toluene separation column is
illustrated in Figure 2.5.  The vent stream from the reflux condenser and
accumulator is first passed through a brine chilled-condenser to remove
organics prior to the steam ejector.  The ejector exit stream is passed

-------
through a vacuum condenser to a hotwell.   The hotwell  vent stream is then
passed through a final brine-chilled condenser prior to exhaust to the
atmosphere.
      Samples were collected from the noncondensable stream (Point 8)  prior
to the ejector for organic species analysis.   Sampling for moisture content
was attempted but was not successful because  of the vacuum present at this
location.
      Samples were collected for organic  species analysis, moisture content,
and inert gas analysis at the hotwell vent before and after (Points 9 and 10,
respectively) the final condenser.  Flow  rate measurements were performed
using a vane anemometer at the vent to the atmosphere (Point 10).  Because
of the low flow rate and low pressure drop across the condenser, it was not
possible to sample simutaneously at Points 9  and 10.  Therefore, these were
sampled sequentially.
      The results of sampling are summarized  in Tables 2.8 through 2.10.
There were significant differences in the results for Cs, 64, and C$ between
runs at Points 8, 9, and 10.  The analysis records were checked to verify
the results.  One reason for the differences  may be that the runs were
sampled on different days.

-------
         From
         Other
         Equipment
             '••'-  :  ; Vent
Overheads  ! •!-  ;
            1   :
OV
       c
Ref 1ux
Condenser
            C
                                                          Decanter Vent
                                   Reflux    1
                                                              Reflux
                                                              Decanter
                                                      Water
                   Benzene Drv-ijig
                     Column       :
                                    ;-...; .':  f"-Q  ' Test  Location
                                                                                                Vent.To    ;
                                                                                              J. Flare Manifold
                                                                                                          cw
                                                    Condensate

                                                      Vent Condenser
                               FIGURE 2.1.  'Benzene  Drying  Column  Vent Equipment.

-------
TO SAMPLE
LOCATIONS
                                                                                              ;Teflon'Sample Line  To

                                                                                               Rigid Container System-

                                                                                                ©
                                                                          GREENBURG-SMITH
                                                                            IMPINGERS
                                                          FIGURE  2.2.   Modified Moisture Train.

-------
    Vent To Flare Mam'foLc
         Caustic Feed
         :(Batfch;
  From Other
  Equipment
  (Off During Test)
            Jest Location
                                   From
                                   Settling  Tank
                                   Caustic  Scrubber
                                   (ECM-002)
                                                    Degasser Vent
                                                    Scrubber
                                                    (ECM-001)
                            Caustic  Drain
                            (Batch Empty)
Figure
2.3. •  Caustic scrubber equipment.  .  ;   :  ;
  •     .'  "          i    '   •        '
                                                                    Vent
                                                                Degasser
                                                                (AS-1008)

-------
                 Vent
                                                 Vent
Catalyst
Storage
  Catalyst Mix
  Tank
                                Catalyst  Mix
                                Tank  Scrubber
                                (ECM-003)
                                        j Drain
                                      To Sewer
                             Return From
                             Scrubber
                         Catalyst Mix
                                                              A Vent To
                                                           ^-'Atmosphere
                                                      Drain To Sewer
                                                      (Capped During Test)
                                                        I Polyethy:! benzene
                                                         • Feed
f-0  :
Test Location
  Figure  2.4.  Catalyst mix tank equipment.

-------
                                                         15015 Steam
                                 Noncondensables
           iCW
   Qverheads ^
           1
      Reflux
                                                            Steam
                                                            Ejector
                   Condensibles
                        Accumulator5
                        Vent Condenser
                                      Brine
Reflux  .
Condenser &
Accumulator
B/T Col umn
f— O   •   Test Location
                                                   Vacuum Condense?
                                                   To Main
                                                   Hotwell
                                                               CW
                                                                            iBrine T
                                                                      HotweU*
                                                                             ^
                                                                                    i Vent  To
                                                                                     'Atmosphere
                                                                     Hotwell Vent
                                                                     Condenser
                                                                     (ECM-005)
                                                                             Condensate
                                                               Hotwell No. 1
  Figure 2.5.  Benzene/toluene column vacuum equipment.

-------
RUN NO.
DATE
TIME
SPECIES ANALYSIS
C-l
C-l
C-2
C-2
C-3
C-3
C-4
C-4
C-5
C-5
C-6
C-6
BENZENE
ETHYLBENZENE
TOTAL HYDROCARBONS
BY:
SPECIES SUMMATION
TOTAL HC ANALYZER
INERTS AND FLOW DATA
HgO, % by volume
N2, % by volume
02 > X by volume
C0£f % by volume
Flow rate, scfm,*dry
HC1 , ppmv
TOTAL
n/m = not measured
* = 20°C atm. , wet :
1A
7/13/78
1045-1055
ppmv as ppmv as
compound benzene
11,025 2,613

92,828 41,257

3,740 2,413

348,970 266,389
840,306 641,455
9,552 9,552

23,808 28,564

2«,185 28,185

1,358,414 1,020,428
n/m
0 0
n/m


IB
7/13/78
1203-1213
ppmv as ppmv as
compound benzene
8,925 2,155

94,160 41,849

3,857 2,488

381 ,957 291 ,570
854,196 652,058
7,671 7,671

30,087 36,097

13,529 13,529
18.71 12,86
1,394,401 1,047,389.86
n/m
0 0
n/m


1C
7/13/78
1437-1447
ppmv as ppmv as
compound benzene
7,350 1,742

77,040 34,240

3,390 2,187

347,230 265,061
774,332 591,093
9,408 9,408

42,645 51,164

28,185 28,185

1,289,580 983,080
n/m
0 0
n/m


Table 2.1.  Results of°Gas-Sampling Analysis-Test Point #1
            Vent Condenser Inlet
                         11

-------
RUN NO.
DATE
TIME
SPECIES ANALYSIS
C-l
C-l
C-2
C-2
C-3
C-3
C-4
C-4
C-5
C-5
C-6
C-6
BENZENE
ETHYLBENZENE
TOTAL HYDROCARBONS
BY:
SPECIES SUMMATION
TOTAL HC ANALYZER
INERTS AND FLOW DATA
H20, % by volume
NZ. * by volume
02 . * by volume
C(>2, % by volume
Flow rate, scfm.*dry
HC1, ppmv
TOTAL
n/m = not measured
* = 20°C atm. , wet
2-1
7/13/78
1045-1055
ppmv as ppmv as
compound benzene
73,500 17,417

194,027 86,234

2,104 1,357

149,377 114,028
291,677 222,654
2,316 2,316

17,660 21,188
16,911 16,911
747,572 482,105
n/m
n/m n/m


2-2
7/13/78
1203-1213
ppmv as ppmv as
compound benzene
. 78,750 18,661

178,809 79,750

2,992 1,885

164,936 125,905
307,302 234,582
50,657 50,657

746,935 896,143
23,675 23,675
1,554,056 1,430,979
n/m
n/m n/m


2-3
7/13/78
1437-1447
ppmv as ppmv as
compound benzene
68,250 16,173

155,982 69,325

2,571 1,659

147,574 112,652
281,260 214,702
2,460 2,461

37,085 44,493
23,675 23,675
578,857 485,140
n/m
n/m n/m


Table 2.2  Results of Gas  Sampling  Analysis- Test  Point #2
                   Vent Condenser Inlet
                             12

-------
LOCATION
RUN NO.
COMPOUND
C5
BENZENE
TOLUENE
ETHYLBENZENE
STYRENE
VOLUME OF
CONDENSATE (ml)
GAS VOLUME
SAMPLED (liters)
CONDENSER INLET
1
9.7
90.3
<0.1
<0.1
<0.1
95
10.0
2
WEIGHT %
8.3
91.7
<0.1
<0.1
<0.1
92.5
10.0
(POINT 1)
3
12.9
87.1
<0.1
<0.1
<0.1
82
10.0
CONDENSER OUTLET (POINT 2)
1 2
WEIGHT %
(INSUFFICIENT



3 1.5
10.0 10.0
3

SAMPLE)



1.5 '
10.0
Table 2.3.  Summary of Results of Species  Analysis-  Organic
              Condensate at Points 1  and 2.
                              13

-------
RUN NO.
DATE
TIME
SPECIES ANALYSIS
C-1
C-l
C-2
C-2
C-3
C-3
C-4
C-4
C-5
C-5
C-6
C-6
BENZENE
ETHYLBENZENE
TOTAL HYDROCARBONS
BY:
SPECIES SUMMATION
TOTAL HC ANALYZER

INERTS AND FLOW DATA
HgO, % by volume
NZ» % by volume
02 t % by volume
C02> % by volume
Flow rate, scfm,*dry
HC1 , ppmv
TOTAL
n/m = not measured
* = 20°C atm., wet
4-1
6/28/78
1045-1100
ppmv as ppmv as
compound benzene
8,099 1,919

8,300 3,689

--

504 385
3,454 2,637
5,970 5,970

252,188 302,565

59,601 59,601

338,116 376,766
n/m

19.69
42.6
7.12
0
53.0
69.41

4-2
7/11/78
inn-inafi
ppmv as ppmv as
compound benzene
8,050 1,908

11,355 5,047

__

5,546 4,234
64,646 49,348
4,713 4,713

64,527 77,417

14,374 14,374

173,211 157,041
n/m

22.77
42.6
7.78
0
53.7
73.15

AVERAGE
ppmv as ppmv as
compound benzene
8,075 1,914

9,828 . 4,382

-r-

3,025 2,310
34,050 25,993
5,342 5,342

158,358 189,991

36,988 36,988

255,666 266,920
n/m

21.23
42.6
7.45
0
53.4
71.28

Table 2.4.  Results .of.Gas Sampling Analysis - Test Point #4
                    Caustic Scrubber.Inlet
                             14

-------
RUN NO.
DATE
TIME

SPECIES ANALYSIS
C-l
C-l
C-2
C-2
C-3
C-3
C-4
C-4
C-5
C-5
C-6
C-6
BENZENE
ETHYLBENZENE
TOTAL HYDROCARBONS
BY:
SPECIES SUMMATION
TOTAL HC ANALYZER
INERTS AND FLOW DATA
H20, X by volume
N2» X by volume
02> X by volume
C02> X by volume
Flow rate, scfm*
HC1, ppmv
TOTAL
n/m = not measured
* = 20°C atm. , wet
5-1
6/28/78
1045-1100
ppmv as ppmv as
compound benzene
288 68

4,427 1,968

--

300 229
1,956 1,493
3,881 3,881

48,012 57,603

56,670 56,670



115i534 121,912
n/m n/m

17.3
43.2
3.08

176.1
0
63.58


5-2
7/1 1 / 7 A
1031 - 1046
ppmv as ppmv as
compound benzene
3.325 788

4,367 1,941

._

2,377 1,814
8,715 6,653
2.062 2,062

42,284 50,731

9,856 9,856



72,986 73,845
n/m . n/m

18.2
n/m
n/m

176.2
0
--


AVERAGE


ppmv as ppmv as
compound benzene
1 ,806 428

4,397 1,955

~

1,338 1,022
5,336 4,073
2,972 2,972

45,148 54,167

33,263 33,263



94,260 97,860
n/m n/m

17.75
43.2
3.08

176.2
0
64.03


Table 2.5.  Results of Gas Sampling Analysis  -  Test Point  £5
          :>         Caustic Scrubber Inlet
                              15

-------
RUN NO.

DATE/TIME
COMPOUND
(ORGAN ICS)
TOTAL
Na+,mg/ml
TOTAL
m-ppm by
Cl vv weight
1
6/28/78


161
34,000
2
6/28/78


159

3
6/28/78


bRun

43
6/28/78
BLANK RUN

0.03

<*Run 4 was used as a blank and calculated into the results,
"Insufficient sample collection.
TABLE 2.6.  Summary of Results of .Caustic Scrubber Liquid
            Analysis Test Point 6.
                         16

-------
RUN NO.
DATE
TIME
SPECIES ANALYSIS
C-l
C-l
C-2
C-2
C-3
C-3
C-4
C-4
C-5
C-5
C-6
C-6
BENZENE
ETHYLBENZENE
TOTAL HYDROCARBONS
BY_:
SPECIES SUMMATION
TOTAL HC ANALYZER

INERTS AND FLOW DATA
t^O, % by volume
N2» t> by volume
02 . % by volume
C02» % by volume
Flow rate, scfm*
HC1 , ppmv
TOTAL
n/m = not measured
* = 2QQC atm. , wet
7-1
6/21/78
1005-1020
ppmv as ppmv as
compound benzene
49.4 11.7

7.2 3.2
60.1 26.7
2.5 1.6

.70 .53





33.2 33.2
153.1 76.93

n/m
7.4
70.4


7-2
6/21/78
1030-1050
ppmv as ppmv as
compound benzene
27.8 6.59

7.0 3.11
65.5 29.11
1.2 ,77

.3 .23





57.2 57,2
159.0 97.01

n/m
7.0
74.2


7-3
6/22/78
inm.iT|i
ppmv as ppmv as
compound benzene
17,0 4.03

4.6 2.04
43.7 19.42
2.3 1.48







7-0 7,0
74.6 33.97

n/m
6.6
68.7


Table 2.7.
Results of Gas Sampling Analysis  -  Test  Point  #7
        Catalyst Mix Tank'Vent
                              17

-------
RUN NO.
DATE
TIME
SPECIES ANALYSIS
C-l
C-l
C-2
C-2
C-3
C-3
C-4
C-4
C-5
C-5
C-6
C-6
BENZENE
ITTLJVI DCU7CUC
tlnlLbtNZtNL
TOTAL HYDROCARBONS
BY:
SPECIES SUMMATION
TOTAL HC ANALYZER

INERTS AND FLOW DATA
HgO, % by volume
NZ» * by volume
02, 2 by volume
CC>2, % by volume
Flow rate, scfin,*dry
HC1 , ppmv
TOTAL
n/m = not measured
* = 20°C atm. , wet
8-1
7/13/78
1202-1213
ppmv as ppmv as
compound benzene
78,750 18,661

34,240 15,218
6,848 3,044
18,701 12,065

15,626 11,928

6,513 6,513



54,114 54,114


214,792 121,543



n/m
n/m


8-2
7/13/78
1630-1631
ppmv as ppmv as
compound benzene
78,750 18,661

45,683 20,290
7,609 3,323
7,692 4,931

25,653 20,277
108 83
4,523 4,523

14,438 17,322

62,006 62,006


246,462 151,416



n/m
n/m


8-3
7/13/78
1654-1655
ppmv as ppmv as
compound benzene
115,500 27,370

50,409 22,404
9,987 4,439
22,675 14,629

32,466 24,783
174 132
5,138 5,138

103,341 123,984

36,076 36,076


375,766 258,955



n/m
n/m


Table 2.8.  Results of.Gas Sampling Analysis  -  Test Point. #8
               Noncondensables to Steam Ejector
                              18

-------
RUN NO.
DATE
TIME
SPECIES ANALYSIS
C-l
C-l
C-2
C-2
C-3
C-3
C-4
C-4
C-5
C-5
C-6
C-6
BENZENE
ETHYLBENZENE
TOTAL HYDROCARBONS
BY:
SPECIES SUMMATION
TOTAL HC ANALYZER
INERTS AND FLOW DATA
H20, % by volume
N2> X by volume
Q£ » X by volume
C0£t % by volume
Flow rate, scfm,*dry
HC1 , ppmv
TOTAL
n/m = not measured
* = 20°C atin. , wet
9-1
6/22/78
1631-1646
ppmv as ppmv as
compound benzene
34,528 8,182.

45,756 20,336
14,536 6,460
2,390 1,542

33,585 25,637

16,000 •• 16,000

85,245 102,273

27,245 27,245

259,285 207,675

28.41
29.0
6.25
21.45
n/m
85.11

' 9-2
6/23/78
0950-1005
ppmv as ppmv as
compound benzene
30,163 7,148

39,840 17,707
11,952 5,312
17,630 11,374

26,046 19,882

10,880 10,880

63,377 76,037

30,439 30,439

230,327 178,779

20.95
20.6
2.68
25.2
n/m
69.43

9-3
6/26/78
1058-1115
ppmv as ppmv as
compound benzene
34,630 8,206

37,840 16,818
8,632 3,836
22,766 14,688

39,196 29,921

9,355 9,355



24,427 24,427

176,846 107,251

57.12
n/m


Table 2.9  Results of Gas Sampling Analysis  -  Test  Point  #9
               Inlet of Hotwell  Vent Condenser  -
                               19

-------

RUN NO.
5-1
5-2
10-1
10-2
10-3
7-1
7-2
7-3
AVERAGE
FT/MIN
2,123
2,083
137.0
86.7
99.3
87.2
82.4
78.4
Ts
95
84
80
70
75
85
85
85
AREA
ANEMOMETER
(F2)
.0873
.0873
.0873
.0873
.0873
.0873
.0873
.0873
FLOW
ACFM
185.3
181.8
12.0
7.6
8.7
7.6
7.2
6.8
SCFM
20°C, 1 atm
176.1
176.2
11.6
7.6
8.7
7.4
7.0
6.6
DSCFM
145.7
144.2
9.6
6.4
7.1



Table 2.11.  Gas Flow Calculations  Test  Points  5,  7,  10.
                          20

-------
RUN NO.
DATE
TIME
SPECIES ANALYSIS
C-l
C-l
C-2
C-2
C-3
C-3
C-4
C-4
C-5
C-5
C-6
f.-6
BENZENE
ETHYLBENZENE
TOTAL HYDROCARBONS
BY:
SPECIES SUMMATION
TOTAL HC ANALYZER
INERTS AND FLOW DATA
HgO, % by volume
N2» I by volume
02 t % by volume
COg. % by volume
Flow rate, scfm *
HC1 , ppmv
TOTAL
n/m = not measured
* = 20°C atm. , wet
10-1
6/22/78
1631-1646
ppmv as ppmv as
compound benzene
37,667 . 8,926

46,849 20,822
13,998 6,221
2,410 1,555

36,049 27,518

15,543 15,543

37,374 44,840
21 ,608 21 ,608

211,498 147,033

16.07
19.7
5.54
18
11.6
59.31

10-2
6/26/78
0920-0935
ppmv as ppmv as
compound benzene
29,002 6,824

39,176 17,412
10,624 4,722
17,630 11,374

25,507 19,471

4,960 4,960

73,242 87,873
32,694 32,694

232,835 185,330

17.82
35.3
8.12
19.5
7.6
80.74

10-3
6/26/78
1015-1040
ppmv as ppmv as
compound benzene
26,880 6,370

29,216 12,985
7,968 3,541
16,126 10,404

24,547 18,738

5,748 5,748

No Data No Data
No Data No Data

110,485 57,786

19.1
n/m
n/m
n/m
8.7
. 76.89

Table 2.10  Results of Gas -Sampling Analysis - Test Point #9
              Outlet from Hotwell-Vent Condenser
                               21

-------
                    3.0 PROCESS DESCRIPTION

      The Cos-Mar plant has two integrated ethylbenzene/styrene units
producing ethylbenzene by benzene alkylation and styrene by ethylbenzene
dehydrogenation.  Both units use the conventional process technology
described in available literature.  The testing was performed on the No. 2
unit, which had a rated capacity of 700(1.0)6 ib/yr styrene.  During all
periods of actual testing, the No. 2 unit was operating between 90 and 100
percent of its capacity.
                                 22

-------
                4.0 LOCATION OF SAMPLING POINTS

      The locations from which samples were collected in each system tested
have been identified in Chapter 2.   In this section, the physical  config-
uration of each location is described and the general stream characteristics
are discussed.  The location identification numbers correspond to  those
referenced throughout this report.
4.1  BENZENE DRYING COLUMN VENT EQUIPMENT
     (Points 1 and 2 on Figure 2.1)
      Both points on the benzene drying column were at elevated pressure.
Point 1 was approximately 90 PSIG and Point 2 was 63 PSIG.   All pipe was
schedule 80 pipe.  The pressure of the stream was used to push the sample
throught the sampling apparatus.  No pump was needed.  As can be seen from
Figure 4.1, the plant had installed a flow measuring orifice.in the pipe
before Point 2.  This orifice was ineffective due to the entrained liquid
discussed earlier.
      Prior to sampling, the residual liquid in both Points 1  and  2 was
purged for a minute or two to purge accumulated liquids from the sample
taps.  The sample line was then connected and the sample was taken.
4.2  CAUSTIC SCRUBBER EQUIPMENT
     (Points 4, 5, and 6 on Figure 2.3)
      Points 4 and 5 represent the inlet and outlet of the caustic scrubber.
Both.Points 4 and -5 were under slight positive pressure.  Normally, the.
scrubber outlet stream is routed to the plant flare system; however, for
testing, this stream was diverted through an existing vent to atmosphere.
A sampling line was connected directly to Point 4 while it was inserted into
the 2-inch diameter hole for Point 5 (Figure 4.2).  Point 6 was a  liquid
sample point.
                                 23

-------
4.3 CATALYST MIX TANK EQUIPMENT
    (Point 7 on Figure 2.4)

      Point 7 was the vent for the catalyst mix tank.   The vent was at the
70-foot level and is normally exhausted 5 feet in the  air from the platform.
At TRW's request, the vent was manipulated so the vent end was accessible
to the samplers on the platform.   The pipe was 4 inches in diameter and was
sampled by inserting the sampling line into the open pipe (Figure 4.3).
Before sampling could occur, the  line below (drain to  sewer)  had to be
closed to force the flow up the vent pipe.
4.4 BENZENE/TOLUENE COLUMN VACUUM EQUIPMENT
    (Points 3, 9, 10 on Figure 2.5)
      Point 8 was on the noncondensable line and was normally under a vacuum
of 27 inches of mercury.  An existing valve was used to extract the sample.
      Points 9 and 10 were at essentially barometric pressure.  An existing
valve was used to extract a sample at Point 9, and the sample line was
inserted into the open pipe at Point 10.   See Figure 4.4 for  the equipment
setup at Points 8, 9, and 10.
                               24

-------
1/40'  Above Ground
   Reflux &
   Decanter Vent
     30' Level ..
                                                                          Teflon Sample Line .
                                                           2"  Di.a.

     To Flare
                                         Teflon Sample Line

                                          25
From
Vent
Condenser
                 FIGURE 4.1--   Benzene drying  column vent equipment

-------
0
85'  Level
   From
   Scrubber
   Degasser Vent
b=o
                               16"
                               3" Dia.
                                                   Valve
                                                                     Sample Line
                                                                     V Teflon
                       \ ...To Caustic Scrubber
Pipe!	[
                    ••Sample
                    Line
                          Valve
         Liquid Sample Valve
                                   From Caustic
                                   Scrubber
                                                  To Air
                                       Caustic Drain
                                            26
                                                                        Anemometer
                                                        Caustic
                                                        Scrubber
                       FIGURE 4.2  Caustic'scrubber equipment.

-------
                     Sample Line
  70'
Ground
                                4"  Pice
                             A»«»'""M>t*
                                              To Air
iFrom Catalyst
 Mix Tank
                                                                       Anemometer
                                              27
                        FIGURE 4.3..  .Catalyst-mix tank  equipment.

-------
                       Valve
emp    L
[idicator
               t
             From
             Hotwel1
          If
         j   Sample
         /    Line
       Y^? "Anemometer
                                                            Sample Line
3"
27"

Did.
Vacuum . "'


                  Jo Hotwell
                   Condenser
                               Vent  To
                               Atmosphere
                              - 2" Dia.
Brine Inlet
                                                                       1" Pipe
                                                        Hotwell Vent
                                                        Condenser
                                                        (ECM-005-V>
                              FIGURE 4.4. Benzene/toluene column  vacuum" equipment.

                                                 28

-------
                   5.0  SAMPLING AND ANALYSIS PROCEDURE

5.1  BAG SAMPLING AND ORGANIC SPECIES ANALYSIS
   The sampling system used for collection of a gas sample into a flexible
container is shown in Figure 5.1.  In all  cases the flexible bag was placed
inside a rigidly sealed container.  The methods used to fill the bags
depended on the pressure of the source gas.  At Points 1  and 2, bags were
filled by source pressure.  At Points 4, 5, 7, 9, and 10, an explosion-proof
pump was used to evacuate the rigid container to about 20" Hg vacuum.  At
Point 8, the rigid container was evacuated to approximately 29" Hg vacuum
prior to sampling.  The container was evacuated and leak checked before
being placed at the test site.  Once at the test site, the Teflon sampling
lines were connected to the sample points, and the values were opened slightly
to produce the desired sampling rate (2 liters per minute).  Once a sufficient
sample was extracted, the container was removed to the laboratory where it
was immediately given to the analyst for processing on the gas chromatograph.
Sampling rates were set based on flow valve setting calibrations prior to
testing.  In some cases, the variability in source pressure (absolute) rigid
container flow restriction or imprecision  of initial flow settings prevented
continuation of sampling for the desired interval.  Because of these variations,
the duration of actual sampling varies from run to run.
   In all tests except those at Points 1 and 2, the sample line was connected
directly to the valve or open pipe that was to be sampled.  At Points 1 and 2,
the bag sampling systems were connected to the outlet of a condenser train
where a side stream was removed.
                                   29

-------
      The bag was run on four separate gas chromotagraphs;
      1) Benzene and higher molecular weight hydrocarbons.
         -Dual FID Shimadzu GC Mini 1 with 5 percent OV-101  and Bentone
          34 - 100/120 mesh
      2) Total hydrocarbons
         -AID Portable FID GC with no column
      3) Cj - C6 (Low molecular weight HC).
         -Dual FID Shimadzu GC Mini 1 with a poropak Q column.
      4) Stationary Gases - 02, N2, C02, CO
         -Shimadzu 3BT Dual Thermal Conductivity with a Chromosorb 102 and
          molecular seive 5x columns.
      The gases were run on these instruments consecutively and the results
were recorded on a linear recorder with a integrator.  In all  cases, the
analysis was done within 4 hours of sampling.
      The audit gases were supplied by EPA and were the only standards avail-
able during the first week of testing.  After the first week,  standards "for
benzene and toluene were shipped to the site.  As can be seen  from the audit
gas report, the concentrations of the audit gases and the standards were
very close.  The ethyl benzene and styrene standards were not available during
the test period.  Liquid samples were used to determine the retention time.
This provided information as to the presence of ethyl benzene and styrene but
was not definitive as to the amount.
5.2 MOISTURE
      A condenser train (Figure 5.2) was used to obtain moisture results.
The procedures used are described in EPA Method 4 (40CFR60 Appendix B)
except that larger impingers were used, followed by a silica gel absorbent.
Concurrent with collection of the water in the samples, organics were condensed
as evidenced by the formation of a lighter liquid layer over the water in the
impingers.  The two liquid layers were separated, with the water phase being
used for moisture content in the source stream and the organic layer being
retained for analysis.  At Sample Locations 4, 5, 7, 8, 9, and 10 the organics
collected were analyzed and reported in Appendix C, but are not reported in
the summary of results because the organics were included in the integrated
bag sample results.
      At Points 1 and 2, the bag samples were extracted after the condenser,
and thus the collected organic condensate is a portion of the  sample.  The
contribution of the liquids  is summarized in Chapter 2.
                                 30

-------
5.3  HYDROGEN CHLORIDE
     HC1 was determined at Points 4, 5, and 7 using IN NaOH as a collection
media and a silver nitrate/potassium thiocyonate titration analysis.  The
complete procedure is included in Appendix E.  During collection of the
sample, using a modified EPA Method 6 train, organics were condensed.  Analysis
for HC1 was performed on the water fraction after separation of the phases.
No attempt was made to analyze the organic layer for HC1.  To determine if
dissolved organics in the water would interfere with HC1 determination,
benzene was added to blank solutions and analyzed.  No interference was
observed.
5.4  ORGANIC LIQUID SAMPLE ANALYSIS
     The samples were run an1 a Shimadzu 6AM gas chromatograph equipped with dual'
column FID with a two-meter glass column packed with 3-percent OV-101 on 80/
100 mesh Gas Chrom R.  A linear integrating recorder was used to record data.
     Standards of the aromatics of interest were run in the concentration
range of 1  percent, 10 percent, and 50 percent in pentane.   Each concentration
is run at an appropriate range and sensitivity of .32V and  103 for the 50
percent.   The resulting integrated area for each standard in each concentration
range was obtained from the chart recorder.   The integrated area for each
aromatic of interest for each sample was obtained.   A linear relationship
between area and concentration is assumed for each concentration range (0.1 to
4 percent), (4 to 20 percent), (20 to 100 percent)  and the  following relation-
ship used to calculate the unknown concentration:
           .  .  .    ...      /<*\ = Area of Unk x Cone, of Std.
     Concentration of unknown (%)   	Area of Std.	

     The samples were stored for approximately four months prior to analysis.
A long sample holding time and room temperature storage introduces several
uncertainties into the analysis for aromatic compounds.
     Aromatics are only slightly soluble in water, so aqueous samples held  in
nongas-^tight  containers at ambient temperatures are subject to possible
losses.  Similar problems are possible  in the nonaqueous samples since losses
of components are a function of their vapor pressures at ambient conditions.
Thus, losses are more pronounced in the lower boiling constituents, making
relative concentrations change in the total sample.
     Chemical changes, especially in unsaturated compounds like styrene, can
result in oxidation.  This can further  affect relative concentrations since
the oxidation products are not extracted as efficiently in extraction steps.

                                    31

-------
     Samples such as those from Point 8 containing less than 5 milliliters
are particularly susceptible to headspace losses because the large gas-to-
liquid ratio in the container allows saturation of a larger gas volume with
vapor.  This could change the relative concentrations in a small  liquid volume.
5.5  CAUSTIC SOLUTION SAMPLE ANALYSIS
     The liquid collected at Point 6 was analyzed by the procedure described
above for organics content.
     Total  sodium and chloride were determined using atomic absorbtion.
5.6  VOLUMETRIC FLOW RATE
     A Rochester G-694 vane anemometer was used to determine the velocity of
the vent streams at Points 5, 7, and 10.  Since the pipes were smaller in
diameter than the anemometer, an expansion adaptor was fabricated.  (The area
of the anemometer face was used to calculate volumetric flow rate.)  The
calibration data show that the length of the expansion section was not enough
to have fully developed flow at the anemometer; therefore, a correction was
used.  The anemometer was calibrated using this configuration.  The calibra-
tion apparatus and results are presented in Figure 5.6.1.  The gas flow calcu-
lations are presented in Appendix A.
                                   32

-------
                                         COARSE
                                        SAMPLING
                                          VALVE
          PROBE
co
GO

                           FLOW
                          METER
                                                            QUICK
                                                             DISCONNECT
                                                              CONNECTORS
QUICK DISCONNECT
 CONNECTORS
                                                                                             VACUUM
                                                                                             GAUGE
                                  FIGURE 5.1.  Modified gas- sampling apparatus.

-------
 TO SAMPLE
 LOCATIONS
co
                                                                            GREENBURG-SMITH
                                                                              IMPINGERS
                               FIGURE 5.2.  Moisture train

-------
oo
en
                                                          FIGURE 5.3.  Anemometer Calibration Apparatus With
                                                                       Correction Results.


                                                          aFt/min Adjusted from Standard Pitot  Readings.

                                                          See Anemometer check with calculations of 1 7/8"
                                                          Readinq and 4" reading to the Anemometer Flows.
                                                          (Appendix  A)

-------