United States
Environmental Protection
Agency
Office of Air Quality
Planning and Standards
Research Triangle Park NC 27711
EMB Report 79-CKO-19
April 1979
Air
Benzene
Coke Ovens

Analysis of Coke Oven
By-Product Process
Samples

-------
           ANALYSIS OF COKE OVEN
         BY-PRODUCT PROCESS SAMPLES
               NOVEMBER 1979
               Prepared for:
        Emission Measurements Branch
    U.S. Environmental Protection Agency
      Research Triangle Park, NC 27711
                    By:
                  TRW
        ENVIRONMENTAL ENGINEERING DIVISION
              P. 0. Box 13000
Research Triangle Park, North Carolina  27709

-------
                              TABLES

NUMBER                                                       PAGE
  1       Benzene and BaP Analytical Results                   2
  2       Partial Pressure Results                             6
  3       GC/MS Results                                        11
                                n

-------
                            CONTENTS
Tables                                              ii
Introduction                                        1
Benzene Analysis Procedure                          1
Benzo-a-Pyrene Analysis Procedure                   13
Appendix                                            16
                               m

-------
                          INTRODUCTION

      Analyses of samples collected at several coke by-products plants were
 performed  in order to provide data for coke oven by-product plants.  Results
 of  these analyses are found in Tables 1 and 2.  As can be seen, the procedure
 for benzene analysis included three phases:  an evaluation of the liquid
 sample and two gas samples.  The first gas phase test is composed of the
 vapor above the sample at the temperature the sample was stored.  The second
 gas phase  test is composed of the vapor above the sample with the storage
 temperature elevated by 20°C.

                   BENZENE ANALYSIS PROCEDURE

      The analytical procedure arose from incorporating comparison analysis
with  samples run on the GC/MS and quality assurance measures.  The instru-
mentation  used initially was found to be extremely effective and accurate.
This was two Shimadzu flame ionization gas chromatographs in conjunction
with  a Shimadzu chromatopak.  The chromatopak was programmed to calculate
peak  area  and retention times which greatly increase the accuracy of measure-
ment  and identification.
      Initially, the liquid samples were analyzed using 5% OV-101 + 1.75% Ben-
tone  34 columns.  The results obtained with these columns proved unsatis-
factory in relation to the estimates of expected concentrations.  The primary
cause of the unsatisfactory results was high concentrations of.substances
with  retention times near that of benzene which caused the benzene peak to be
obscured.                                ,
     Since the GC/MS values corresponded well with the expected concentrations,
another set of samples were analyzed using the same chromatographic conditions
                                                                             I
                                                                         \
                               1

-------
                                       TABLE  1.  BENZENE AND BoP ANALYTICAL RESULTS

SAMPLE
NUMBER
S77-001-601
S77-001-603
S77-001-604
S77-001-605
S77-001-606
S77-001-607
S77-001-608
1X3 S77-001-609
S77-001-610
S77-001-611
S77-001-612
S77-001-613
S77-001-614
S77-001-615
S77-001-616
S77-001-618
S77-001-620
S77-001-621
S77-001-622
. S77-001-623
S77-001-633
S77-001-634
TYPE OF SAMPLE
Treated Wastewater
Untreated Wastewater -
20% skim, 80% water sample
Water
Napthalene . .
Napthalene
Water
Untreated Wastewater -
Skimmed Oil Sheet
Wastewater
Coal Tar
Weak Ammonia Water
Water—Organic Layer
Debenzolyzed Wash Oil
Light Oil — High Crack
Light Oil -- Medium Crack
Separated Water
Tar and Condensate
Tar
Weak Ammonia Water
Weak Ammonia" Water
Tar Weak Ammonia Water
Spent Carbon Slurry
Effluent From Flotation Unit
BaP
0.0018
0.0233
0.004
. 1.380
1.130 .
0.0048
15.500 •
0.0015
18.000
0.100
0.0063
43.500
0.170
3.700
0.003
0.120
23.000
0.115
0,170
0.170-
0.0001
0.0092
BENZENE
LIQUID
CONC.
(ppmv)
10
2
4
350
<1
1,224
'
-------

SAMPLE
NUMBER
S77-001-635
S77-001-636
S77-001-637
S77-001-638
.577-001-639
S77-001-640
577-001-641
S77-001-642
1 S77-001-643
S77-001-644
S77-001-645
577-001-646
577-001-647
577-001-649
577-001-651
577-001-653
S77-001-655
577-001-657 •
577-001-659
577-001-661
577-001-663
TYPE OF SAMPLE
Carbon Adsorber Effluent
Intercepting Sump Water --.Stream B
Intercepting Sump Oil .— Stream B
Enriched Final Cooler Oil
Absorber Liquor Circulatory Tank
Tar Storage
Tar Storage Sump
Benzol Scrubber Sump
Ammonia Absorber Effluent
Influent to Wastewater Treatment
Flushing Liquor Decanter - Tar
Tar Storage .
Cold Well Water
Hot Well Water
'Tar Intercepting Sump
Oil Intercepting Sump
Tar From Final Cooler
Contaminated Sump Inlet
•Contaminated Water From Decanter
Oil Layer From Decanter
. Feed to Crude Stills
BaP
<0.0001
28.000
5.000
18.000
0.400
25.500
0.0322
41.500
0.0013
0.011
18.000
12.900
0.0223
O.Q352'
8.500
11.500
0.300
3.500
0.090 . .
. 7.000 .
1,320
BENZENE
LIQUID
"CONC.
(ppmv)
841
480
590
18,320
4
1.951
7
4,2561
2
2
11
6.321'
. 54
157
6
4,532
.101
560,145
29,420
120,251 .
44,383
VAPOR
CONC. TEMP.
(ppmv) (°C)
282
2,007
1,847
3,536
1,281
421
897
8,397
2,552
617
2,821**
1,229
14,427
974
221 '
5,096
18,900
75,069
15,927
5,113
9,851
45
60
60
40
48
66
70
54
33
74
96
62
28
44
49
. 28
28
32
31
17
26
VAPOR
CONC. TEMP.
(ppmv) (°C)
687
4,998**
3,721**
10,590
6,201
1,827**
3,876**
21,682
7,837
3,681**
	 *
4,780**
38,791
3,606
.601
6,793
57,251
' 168,553
. 63,108
. 34,547
64,439
65
80
80
60
68
86
90
74
53
94
116
82
48
64
69
48
.48 .
52
.51
37
46
 *NOT ANALYZABLE BY METHOD DUE TO TEMPERATURE
**VALUES QUESTIONABLE DUE TO BOILING POINT OF BENZENE
Values checked by GC/MS

-------

SAMPLE
NUMBER
577-001-665
S77-001-667
577-001-669
577-001-671
S77-001-673
577-001-675
S77-001-677
S77-001-679
*> 577-001-681
577-001-683
577-001-685
577-001-687
577-001-689
577-001-691
S77-001-693
S77-001-695
S77-001-697
S77-001-699
S77-001-801
S77-001-802
S77-001-803
TYPE OF SAMPLE .
Drainage From Decanter for #23 Still
Drainage From Decanter for #23 Still
#20 Still Decanter Effluent
Methyl Naphthalene in Circulating Tank
#2 Carbolic Oil
#1 Carbolic Oil .
Middle Creosote
Light Creosote
Heavy Creosote
Pitch
Oil /Water Mixture From V113 Tank (oil)
. Oi 1 /Water Mixture From V113 Tank (water)
Methyl Naphthalene From #1 C.O.
Slop Tank Sump
Light Tar Storage Tank #2 .
Heavy Tar Storage Tank. #3
Inlet to Deperator
Final Cooler Sump
Rich MEA Solution
Feed to Denver Units
Debenzolyzed Wash Oil Decanter Oil
BaP
(yg/ml )
0.001
<0.0001
0.080
10.000
 W  W I  I 1^ I I IV V IS V i**  IT*  • MI ti •— • w • • » i i —
**VALUES QUESTIONABLE DUE TO BOILING POINT OF BENZENE
'Values checked.by  GC/MS

-------

RFM7FMP
SAMPLE
NUMBER


577-003-804
S77-003-805
S77-003-806
S77-003-807
TYPE OF SAMPLE


Gale Sump Water Layer
Gale Sump Oil Layer
Underflow From Denver Unit
Overflow Pit/Sump
BaP
(pg/ml }


.140
0.001
.0015
.110

LIQUID
CONC.
(ppmv)
221
1,294
26,5
104,900

VAPOR
CONC.
(ppmv)
1,920
2,140
505
12,380


TEMP.
(°c)
48
26
28
55

VAPOR
CONC. ,!.
(ppmv) ;
4,880
2,998
1,280
47,150


TEMP.
(°c)'
68
46
48
75

-------
                                          TABLE 2.   PARTIAL PRESSURE RESULTS
SAMPLE
. 577-001-601
S77-001-603
S77-001-604
.577-001-605
^'•577-001-606
577-001-607
577-001-608
S77-001-609
577-001-610
577-001-611
577-001-612
577-001-613 •
577-001-614
S77-001-615
•577-001-616 '
577-001-618 .
. $77-001-620
577-001-621
PARTIAL
PRESSURE
(mmHg )
0.0091
0.0053
0.148
0.365
0.130*
0.098
0.00574
••> 0.0845
1.82
1.59*
. - 0.89*
0,129
2,20
4,55—
1.53
"'OJ60
!5;
STORAGE
TEMP. '
21 •
21
38 • •
20
93
18-
29
29
76
82 •'.
26 .
51
. : 25 .
82
58
. -54
45; •
BAROMETRIC
PRESSURE
(mmHg )
758
755 .
765
750 • •
762 •
750'
770
761
750 . •
. 755 • '
755 . ' '" .
' . 761 .
762 . •"••.-•
771 - '
758
770 '
' • ' 764 '
772
PARTIAL
PRESSURE
' (mmHg)
o.on
0.0067
0.2066
1.63
-• **
0.162 .
0.0274 .
0.224
3.63*
.- **
1.560
0,0337
. . 4.34 •
5,41*
' -'•"2t37 .'
. • . '0,831
..'• ij1^
STORAGE
' - TEMP. + 20°C
41
41.
• 58 • .
40
- .113
38 '
49
49.
96. .
' 102
46
71
45
102
78
74 .
..: ,.:,^;- ' ••
BAROMETRIC
. PRESSURE
762
748
738
. 753
„_
754
762
758
755
--
739
748
749
751
762 .
752

LVALUES QUESTIONABLE DUE TO BOILING  POINT-OF BENZENE,
**NOT.ANALYZABLE  BY  METHOD  DUE TO'TEMPERATURE..

-------
                                     TABLE 2.  PARTIAL PRESSURE RESULTS (cont'd)-
SAMPLE
S77-001-622
S77-001-623
S77-001-633
$77-001-634
S77-001-635
S77-001-636
S77-001-637
S77-001-638
S77-001-639 .
S77-001-640
$77-001-641
S77-001-642
S77-001-643
S77-001-.644
S77-001-645
S77-001-646
$77-001-647
S77-001-649
PARTIAL
PRESSURE
(mmHg)
--**
0.291
0.113
0.553
0.217
1.523
1.40
2.68
9.971
0.320
0.685 '
:6.38
1.95. •
•0.470
2.13*
0,930
11,11
0.739
STORAGE
TEMP.
(°c)
100
51
40
46
45
60
60
40 '
48 .
66
70
54
33
'74
96
62
28 .
44
BAROMETRIC
PRESSURE
(mmHg)
-- ** '
765
769
768
770
759 -
760
759
758
760 ..
764
760 .
764
762
755
757 .
770 •' .. . •'
759 '
PARTIAL
PRESSURE-
(mmHg)
— **
: 0.428
0.396
2.02
0.521
3.82*
2.89*
8.09
4.58
. 1.40*
. 2.94*
16.52
5.87
2.79*
.. **
3.59* :
28.9 ;•,-:••
. 2.77
STORAGE
TEMP + 200C
(°C)
120
71
60
66
65
80
80
.60
68
86
90
74 .
53
94
116
•82
48
64
BAROMETRIC
PRESSURE
(mmHg)
—
740
747
770
758
765
• 777
764
739
765
759
762
749
.759
—
' 751
745
768
 *VALUES .QUESTIONABLE DUE TO BOILING POINT OF BENZENE-
**NOT ANALYZABLE  BY  METHOD.DUE TO TEMPERATURE, .

-------
                                     TABLE 2.  PARTIAL PRESSURE RESULTS (cont'd)
SAMPLE
577-001-651
S77-001-653
S77-001-655
S77-001-657
oo 'S77-001-659
S77-001-661
S77-001-663
S77-001-665
S77-001-667
S77-001-669
S77-001-671
.S77-001-673
S77-001-675
S77-001-677 .
577-001-679
S77-001-681
S77-001-683
. 577-001-685
PARTIAL
PRESSURE
(mmHg)
0.168
3.87 •
14.57-
57.20
12.30
3.83
7.60
17.38
17.67
2.09.
10.07
11,41, .
13,52 :.'
3.20*
. 0.130
0.756* .
' 0.168* .
• 39,91'
STORAGE
TEMP.
4.9
28
28
32
31
17
26
21
21
24
65
7.9
42"
150
62
104
23,9
60
BAROMETRIC
PRESSURE
(mmHg)
760
760
771
762
772 .
750
771
768
766
770 . .
758
770 •
756
765
760 ..
771 , ' '
759 '. . ..'
762 •' .
PARTIAL
PRESSURE
(mmHg)
0.450
5.15
43.8
128.1
46.7
26.05
48.84
20.29'
33.15
10.0 .
35.39*
. . 29.11*
33,8
•. — **
0.562
.' ' 2,13
„„** . _
• ' " 149,0*
STORAGE
TEMP + 2QOC
69
,48
48
52
.51
37
. 46
41
41
.44
85
99 .
. , 62
170
82
.124 .
••' ' 259 • '• • .•
. ' 80
BAROMETRIC
PRESSURE
(mmHg )
747
758
765
760
740
754 . •
758
762
764
762 . •
750
.741
748 •
•*••
768
' 762
•' 750' • ; '
 *VALUES QUESTIONABLE DUE TO BOILING POINT OF BENZENE,
**NOT ANALYZABLE BY  METHOD  DUE TO TEMPERATURE,  •

-------
                                TABLE 2.  PARTIAL PRESSURE RESULTS (concluded)
SAMPLE
S77-001-687
S77-001-689
577-001-691
S77-001-693
S77-001-695
S77-001-697
S77-001-699
S77-003-801
S77-003-.802
S77-003-803
S77-003-804
S77-003-805
S77-003-806
S77-003-807
PARTIAL
PRESSURE
(mmHg)
1.55
2.40. .
. 0.561
0.301
4.60
6.47 '
4.79
0.0924
0.678
0.965
1,45 .
1.63 '
0.389
9.5Q- .
STORAGE
tEMP. •
60
60
" 37
60
71
69
52
32
29 '
28
48
26
28-
55
BAROMETRIC
PRESSURE
(mmHg )
771
774
770 •
759
774
768
762
764
758
755
754
761
770
767
PARTIAL
PRESSURE
• (mmHg)
3.19*
6.77*.
1.81
1.24* .
10.92*.
22.27*
14.41
0.307
1.56
3.01
3.72 :
2.25
0.978
• 35.41
STORAGE
TEMP + 20°C
(OC)
80
80 .
57
80
91
89 ,
72
52
49
48
68 .
46
48
75
BAROMETRIC
PRESSURE
(mmHg)
758
757
752
762
768
758
760
780
111
772
762
750
764
751
*VALUES QUESTIONABLE DUE TO BOILING POINT OF BENZENE,

-------
 as-the  GC/MS.  (The GC/MS results are presented in Table 3.)  This procedure
 called  for 0.2%. Carobwax 1500 on Carbopak-C columns with a temperature
 program of 50°C to 140°C at a rate of 16°C per minute.  The procedure
 followed was identical, with the exception of a 15°C per minute rate for the
 temperature program set by the 5°C per minute rate increments on the Shimadzu
 gas  chromatographs.  The 0.2% Carbowax 1500 on Carbopak-C columns were much
 more desirable  for this analysis since they efficiently separated benzene 7
 from the other  substances, regardless of their concentration, by increasing
 the  retention times of each substance.  The second set of samples produced
 the  same values  obtained with the GC/MS, so this procedure was adopted for
 the  rest of the  analysis.
     Tv/o problems became apparent with respect to reproducibility and quality
 in the  analysis  of the liquid samples.  First, a procedure for blanking the
 gas  chromatograph and injection syringes was .needed.   It was found that the
 instrument  required time and flushing with both carbon disulfide and nitrogen
 to remove high concentrations of substances, including benzene, found in many
 of the  liquid samples.   The syringes were also contaminated by each liquid
 injection and flushing with carbon disulfide with the aid of a high vacuum
 pump and syringe cleaner was necessary.   For quality assurance purposes,
 blanks  on both the instrument and syringe were required before each sample
 injection.   This repeatability requirement for each sample increased greatly
 the time required for analysis.
     The second problem encountered with the liquid samples arose in the prepa-
 ration of accurate standards.   Initially, standard pipetting methods and serial
                                         '»
 dilutions were used but proved inaccurate.'  To remedy this situation, a 5 ml
syringe and serial dilutions were used to prepare the standards.  After many
repeatable  runs, these  standards were plotted graphically and mathematically
                                                                           I
                               10

-------
                            TABLE 3



                         GC/MS RESULTS

SAMPLE #
577-001-640
577-001-642
577-001-646
577-001-693
ppmv
by
GC/MS
4.0
>2000*
7.0
8.0
ppmv
by
GC
1.95
4,256
6.32
9.12
* Detector flooded
                               11

-------
and were found to be linear, thus assuring the quality of the standards.
     The gas phases of the analysis also had similar problems.   For both  the
ambient collection temperature tests and the 20°C elevation temperature tests
the Shimadzu Gas Chromatographs with the 0.2% Carbowax 1500 on Carbopak-C
columns and 50°C to 140°C temperature program were used.   Again with the  gas
phases, there were problems associated with blanking the  machines and syringes.
To remove the contamination in the columns, a flushing with pure nitrogen gas
was used.  A switch from glass syringes to disposable syringes removed the
necessity for blanking the syringe between each sample.
     The standards used for the gas analysis were commercially prepared by
Scott Specialty Gases.  Several different concentrations  were analyzed,
plotted graphically and found to be linear, thus assuring quality of the  stan-
dards.
     A third quality assurance measure was taken in the gas phases analysis.
The Shimadzu gas chromatographs have a separate external  loop for gas sampl-
ing which aids in sample size repeatability.  The trap size is 1 ml but a
10 ml  sample is injected through the loop to assure a complete sample. The
sample in this analysis is specific for temperature so the sample loop was
controlled at the gas collection temperature.  This assures that the sample
is monitored at the required temperature.
     One additional  procedure was instituted to insure validity of the results
for the gas phase analysis.   The initial analysis was carried out by storing
the liquid in a container controlled at the liquid storage temperature.  For
the lower temperature samples (up to 40°C), a glass vial  with teflon lid  and
septum was used to store the sample while 'bringing it up to the storage
perature.  For higher storage temperature samples, a standard Parr bomb
apparatus v/as used to eliminate the possibility of container rupture due to
                               12

-------
 internal pressure.   This  device  is  a  completely  sealed steel container with



 an inner container of all  teflon.   In both  cases the containers were carefully



 opened under a hood to release pressure,  then  a  gas phase sample was quickly



 extracted through either  the  septum or a  small opening around the lid using



 a gas-tight syringe.   The  sample was  then raised slowly by 20°C and the



 procedure repeated.   Some  samples showed  lower values at the + 20°C tempera-



 ture,  indicating benzene  depletion  in the sample.  To correct this- problem,



 a fresh aliquot was  used  for  the temperature + 20°C analysis.  Temperatures



 were maintained by  immersing  the sample in  a constant temperature water bath,



 and in all  cases the  liquid level in  the  containers was approximately 75



 percent of  the volume  of  the  container.   The containers were 80 mi Hi liters in



 volume.



     The procedure  described  above  with the modifications for both liquid



 and gas  samples was adopted and all the samples were re-analyzed using these



 methods  and quality assurance techniques.



     Several  of the samples collected for benzene were to be analyzed for



 the vapor phase at  temperatures above the normal boiling temperature for



 benzene. This v/ould  produce  a boiling  condition, which would not give a



 representative, steady state  value.   Therefore, these values are questionable.



 A few  samples could not be analyzed effectively due to the frothing or rapid



 boiling  at  the designated  temperature.                                 •





               BENZO-a-PYRENE  ANALYSIS PROCEDURE    .              '





     Fluorescence spectrophotometry was used to analyze for BaP.  The



 equipment used for  this analysis was  the  Aminco Model SPF-125 Spectrophoto-



 fluorometer with 7mm  lightpath cell.   This  instrument accurately measures



•concentrations of BaP  as  low  as 0.001  ppm.  The wavelength settings were



 378 nm excitation and  403  nm  emission with  respective slitwidth openings of





                               13

-------
 Imn  and  5mm.  All the samples were in a liquid state, so the only preparations


 involved examining each sample and diluting in cyclohexane any sample which


 was  darkly colored, contained abundant suspended material, or was extremely


 viscous.  This was necessary because any particles or opacity will affect


 the  absorbence.  This instrument becomes extremely substance specific at very


 narrow slit widths, as were used in this analysis.


     Filtered particulates and solid samples required an eight hour extrac-


 tion period in cyclohexane before analysis could be performed.  A cyclohexane


 blank was run and taken into account on all extracted and diluted samples in


 the  final calculations.


     The quality assurance procedures taken for this analysis included pre-


 paration of a series of BaP standards, exclusion from light, and spiking.  A


 set of standards were prepared for each range (high, medium, and low concen-


 trations) by serial  dilutions.  Each set was'"analyzed for linearity by


 continual measurement throughout the days testing.  Since BaP is light sensi-


 tive, standards and sample aliquots were discarded after analysis and the


 samples were kept in closed, dark containers.  Lastly, a spiking procedure


was used to determine recovery efficiencies on solid and filtered samples,


and on samples with  very low BaP concentrations.  A spiking procedure was


followed to assure accurate detection near the limits of the machine.


     No major problems were encountered with the fluorescence spectrophoto-


metric procedure for BaP analysis.   This method is preferred over the thin


layer chromatographic (TLC) method for low level BaP analysis, as the TLC


method had only about 0.01 the sensitivity of direct liquid measurement..
                                         r,

     On higher concentration  samples, selected samples were re-run using TLC


and a TLC scanning attachment for the fluorescence spectrophotometer.
                               14

-------
 Comparable results were produced.
     The  benzo-a-pyrene method was tailored to these samples.   The method
 originally chosen was intended to be a thin layer chromatography  separation
 with measurement by scanning in-situ with a scanning attachment for the
 fluorescence spectrophotometer.
     This method lacked the sensitivity required for the analyses.  The
 samples were to be pre-concentrated using Kuderna-Danish concentrators with
 a nitrogen stream flowing over them.  However, most of the samples would not
 reduce in a reasonable time.  Some samples took several weeks  to reduce their
 volume by 50 percent while others would not reduce within two  weeks, even
with ethanol  added.  It was found by reviewing the literature  that no compound
expected to be present in these samples had a similar excitation/emission
spectrum to benzo-a-pyrene.  Previous analyses by GC/MS on similar samples .
were the basis of the compounds to be considered as interferences.  In. addi-
tion, a general compendium of poly-organic materials and other organtcs showed
no similar spectrum.    It was, therefore, decided to go to a direct in-situ
method as previously described.
 Fluorescence and Phosphorescence Data Compendium,
Donald L. Helman, American Instrument Co., 1977.
                               15

-------
                 APPENDIX

Calculations
Benzene
Partial Pressure
Benzo-a-Pyrene
Typical Chormatograms - Benzene Standards
Typical BaP Calibration Chart
Typical BaP Calibration Curve
                    16

-------
                          CALCULATIONS



Benzene



   1.  Prepare standard curve to check linearity.



   2.  Calculate constant (K) by:  area (from chromatopak)
                            '  J   concentration of standard



   3.  Concentration of benzene in sample:  (area) x (K)



Partial Pressure



   Assuming an ideal gas:   ^oles benzene      (pressure)

                           10  moles total


       Therefore:  PPmv benzene     (Atmospheric Pressure mm Hg)

                   10  ppmv total



Benzo-ct-Pyrene



   1.  Prepare a standard  curve.


   2.  Read, "as analyzed" concentration from curve.


   3.  Correct by dilution factor, if any.
                                17

-------
                         TYPICAL CHROMATOGRAM

                          BENZENE STANDARD
  ,'.. .-'--.i PPM
  fIVEL        21.2
  THPT  OR. .16. 15.28.
                9.4
                                                                                    '•ti
      •• r! i H
                     18=74
C-RtH
?:MPL *
FILE *
REPT ft
METHOD
ji
0
iji


fl£l
2
6
41
HHHE TIME
9.4
10.74
'i'OTflL



p
'•^•'
rfiNC f'1K
4.4572
93.5427
99.9999





RRER
319
6838
7158
i EVEL
-15
 2J..4
                                   18
                                                                Vi. -

-------
                       TYPICAL CHROMATOGRAM


                        BENZENE STANDARD
STflRT  Q8.16.15.52.

    fi.4
LEVEL     -281.2
L£V EL
           -18
                                                              16.34
  STOP
ORlfl
SMPL tt       m
FFLE *        2

METHOD       41

    »   HHf'iE   TIME
    e
CONC      MK
 0.3335
i-3.^-4
                                                RRtH
                                                  377
                                                             c a «\ •<
                                 19;

-------
                        TYPICAL CHROMATOGRAM

                          BENZENE STANDARD
     PPM
     "   98,16.16.33.
                                                                 18.41
      STOP
f:-P!R
             00
FILE *
PEPT #
METHOD
if
i?
2
•3
41
MRMF




TIME
e.33



COHC
0.071




9
                                                  flREfl
                                                    376
                                                          , - -_ . c <
                                   20

-------
•F"
5
/L
          O
     1
A_JL
                  ,_..
                                      -
                                                    P xx
                                                    \ -
                  ~i
                     _L_I^_
                           f
                                           V3

                                           ~n~rr^z:~!'~'
                TT
                                              1-1-
                                                     -pJ_r^
                       i,   -i'j:,-
                    BaP STANDARDS - pg/ml
                           21

-------
  Calibration  Curve  for BAP  Analysis
       0-
    *•;.-•/>

A ••'-••-
              /
                                 -          I,
                                 •'  V   -  o
                                 ;   -i.  A  1-
                                 ri   c   •  S>
                                 ^   ^   o |
                                  U.
                                  V!

                                   I
                                                   v °-°
                                                   •i .


                                                   u.'

                                                    I

-------