United States          Environmental Monitoring Systems
               Environmental Protection      Laboratory
               Agency             Research Triangle Park NC 27711

               Research and Developmen     EPA/600/4-77/027b June 1987
f/EPA          Quality Assurance
               Handbook for
               Air Pollution
               Measurement
               Systems:
               Volume III. Stationary
               Sources Specific
               Methods

               Sections 3.0.4, 3.0.7
                          3.0.9, 3.0.10,
                          3.14, and 3.15
               •Note: Sections 3.0.4, 3.0.7, 3.0.9, and 3.0.10 are technical guidelines
                  to support the June 4, 1987, EPA Promulgated Final Rules for
                  Quality Assurance Requirements for Gaseous CEMS used for
                  Compliance Determination (Appendix F to 40 CFR 60).

-------
 June 1987
                             Volume III

                       Table of Contents


Section                                         Pages    Date


            Purpose and Overview of the Quality           3       1 -04-85
            Assurance Handbook

    3.0      General Aspects of Quality Assurance for
            Stationary Source Emissions
            Testing Programs

    3.0.1    Planning the Test Program                   11       5-01-79
    3.0.2    General Factors Involved in Stationary          2       5-01-79
            Source Testing
    3.0.3    Chain-of-Custody Procedure for Source         7       5-01-79
            Sampling
    3.0.4    Procedure for NBS-Traceable Certification     14       6-09-87
            of Compressed Gas Working Standards
            Used for  Calibration and Audit of
            Continuous Source Emission Monitoring
            (Revised  Traceability Protocol No. 1
    3.0.5    Specific Procedures to Assess                56       9-23-85
            Accuracy of Reference Methods Used for
            SPNSS
    3.0.6    Specific Procedures to Assess                14       9-23-85
            Accuracy of Reference Methods Used for
            NESHAP
   X3.0.7    Calculation and Interpretation of              14       11-05-85
            Accuracy for Continuous Emission
            Monitoring Systems (CEMS)
    3.0.8    Audit Materials Available from                7       11-04-85
            U.S.E.P.A.
   ^3.0.9    Continuous Emission Monitoring             47       6-01-86
            Systems (CEMS) Good Operating
            Practices
    3.0.10  Guideline for Developing Quality             11       11-26-85
            Control Procedures for Gaseous
            Continuous Emission Monitoring
            Systems

    3.1      Method 2—Determination of Stack Gas
            Velocity and Volumetric Flow Rate
    3. .1     Procurement of Apparatus and Supplies
    3. .2     Calibration of Apparatus
    3. .3     Presampling  Operations
    3. .4     On-Site Measurements
    3. .5     Postsampling Operations
    3. .6     Calculations
    3. .7     Maintenance
    3. .8     Auditing Procedure
    3.1.9     Recommended Standards for Establishing
             Traceability
    3.1.10   Reference Method
    3.1.11    References
    3.1.12   Data Forms
15
21
7
12
3
4
1
5
1
11
2 1
8 1
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
1-15-80
1-15-80
1-15-80
-15-80
-15-80
-15-80
    3.2      Method 3—Determination of Carbon
             Dioxide, Oxygen Excess Air, and Dry
             Molecular Weight

-------
                                                      June 1987
Section
Table of Contents (continued)

                                     Pages
                                                             Date
15
4
6
12
2
3
1
5
1
3
1
6
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
3.2.1     Procurement of Apparatus and Supplies
3.2.2     Calibration of Apparatus
3.2.3     Presampling  Operations

3.2.4     On-Site Measurements
3.2.5     Postsampling Operations
3.2.6     Calculations
3.2.7     Maintenance
3.2.8     Auditing Procedure
3.2.9     Recommended Standards for
         Establishing Traceability
3.2.10   Reference Method
3.2.11    References
3.2.12   Data Forms

3.3      Method 4—Determination of Moisture
         in Stack Gases

3.3.1     Procurement of Apparatus and Supplies
3.3.2     Calibration of Apparatus
3.3.3     Presampling  Operations
3.3.4     On-Site Measurements
3.3.5     Postsampling Operations
3.3.6     Calculations
3.3.7     Maintenance
3.3.8     Auditing Procedure
3.3.9     Recommended Standards for
         Establishing Traceability
3.3.10   Reference Method
3.3.11    References
3.3.12   Data Forms

3.4      Method 5—Determination of Particulate
         Emissions from Stationary Sources

3.4.1     Procurement of Apparatus and Supplies
3.4.2     Calibration of Apparatus
3.4.3     Presampling  Operations
3.4.4     On-Site Measurements
3.4.5     Postsampling Operations
3.4.6     Calculations
3.4.7     Maintenance
3.4.8     Auditing Procedure
3.4.9     Recommended Standards for
         Establishing Traceability
3.4.10   Reference Method
3.4.11    References
3.4.12   Data Forms

3.5      Method 6—Determination of Sulfur
         Dioxide Emissions from Stationary Sources
   3.5.1     Procurement of Apparatus and Supplies        6        5-01-79
   3.5.2     Calibration of Apparatus                      6        5-01-79
   3.5.3     Presampling Operations                      3        5-01-79
   3.5.4     On-Site Measurements                       7        5-01-79
   3.5.5     Postsampling Operations                     7        5-01-79
   3.5.6     Calculations                                2        5-01-79
   3.5.7     Maintenance                               1        5-01-79
   3.5.8     Auditing Procedure                          6      -  5-01-79
9
19
7
10
4
8
3
4
1
5
1
14
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
-15-80
                                       15
                                       22
                                       20
                                       19
                                       15
                                       10
                                        3
                                        7
                                        1

                                        6
                                        2
                                       21
                                                                1-15-80
                                                                1-15-80
                                                                1-15-80
                                                                1-15-80
                                                                1-15-80
                                                                1-15-80
                                                                1-15-80
                                                                1-15-80
                                                                1-15-80

                                                                1-15-80
                                                                1-15-80
                                                                1-15-80

-------
    June 1987
               Table of Contents (continued)


Section                                          Pages     Date

   3.5.9     Recommended Standards for                  1        5-01-79
            Establishing Traceability
   3.5.10    Reference Method                            4        5-01-79
   3.5.11    References                                  1        5-01-79
   3.5.12    Data Forms                                 13        5-01-79

   3.6       Method 7—-Determination of Nitrogen
            Oxide Emissions from Stationary Sources

   3.6.1     Procurement of Apparatus and Supplies         5        5-01-79
   3.6.2     Calibration of Apparatus                      5        5-01-79
   3.6.3     Presampling Operations                      5        5-01-79
   3.6.4     On-Site Measurements                       8        5-01-79
   3.6.5     Postsampling Operations                      5        5-01-79
   3.6.6     Calculations                                 4        5-01-79
   3.6.7     Maintenance                                1        5-01-79
   3.6.8     Auditing Procedure                           6        9-23-85
   3.6.9     Recommended Standards for                  1        5-01-79
            Establishing Traceability
   3.6.10    Reference Method                            5        5-01-79
   3.6.11    References                                  1        5-01-79
   3.6.12    Data Forms                                 13        5-01-79

   3.7       Method 8—Determination of Sulfuric Mist
            and Sulfur Dioxide Emissions from
            Stationary Sources

   3.7.1     Procurement of Apparatus and Supplies         7        5-01-79
   3.7.2     Calibration of Apparatus                     10        5-01-79
   3.7.3     Presampling Operations                      4        5-01-79
   3.7.4     On-Site Measurements                      10        5-01-79
   3.7.5     Postsampling Operations                      9        5-01-79
   3.7.6     Calculations                                 6        5-01-79
   3.7.7     Maintenance                                2        5-01-79
   3.7.8     Auditing Procedure                           3        5-01-79
   3.7.9     Recommended Standards for                  1        5-01-79
            Establishing Traceability
   3.7.10    Reference Method                            5        5-01-79
   3.7.11    References                                  1        5-01-79
   3.7.12    Data Forms                                 17        5-01-79
   3.8      Method 10—Determination of Carbon
            Monoxide Emissions from Stationary
            Sources

   3.8.1     Procurement of Apparatus and Supplies        13       1-04-82
   3.8.2     Calibration of Apparatus                     18       1-04-82
   3.8.3     Presampling Operations                      6       1-04-82
   3.8.4     On-Site Measurements                      12       1-04-82
   3.8.5     Postsampling Operations                     5       1-04-82
   3.8.6     Calculations                                3       1-04-82
   3.8.7     Maintenance                                2       1-04-82
   3.8.8     Auditing Procedure                          7       1-04-82
   3.8.9     Recommended Standards for                  7       1-04-82
            Establishing Traceability
   3.8.10   Reference Method                           3       1-04-82
   3.8.11    References                                  2       1-04-82
   3.8.12   Data Forms                                11       1-04-82

-------
                                                       June 1987
Section
Table of Contents (continued)

                                    Pages
Date
   3.9      Method 13B—Determination of Total
            Fluoride Emissions from Stationary
            Sources (Specific-Ion Electrode Method)

   3.9.1     Procurement of Apparatus and Supplies        20       1-04-82
   3.9.2     Calibration of Apparatus                      25       1-04-82
   3.9.3     Presampling Operations                       6       1-04-82
   3.9.4     On-Site Measurements                      21       1-04-82
   3.9.5     Postsampling Operations                     19       1-04-82
   3.9.6     Calculations                                 7       1-04-82
   3.9.7     Maintenance                                3       1-04-82
   3.9.8     Auditing Procedure                           8       1-04-82
   3.9.9     Recommended Standards for                   1       1-04-82
            Establishing Traceability
   3.9.10    Reference Method                            2       1-04-82
   3.9.11    References                                  1       1-04-82
   3.9.12    Data Forms                                 22       1-04-82

   3.10     Method 13A—Determination of Total
            Fluoride Emissions from Stationary
            Sources (SPADNS Zirconium Lake
            Method)
   3.10.1    Procurement .of Apparatus and Supplies
   3.10.2    Calibration of Apparatus
   3.10.3    Presampling Operations
   3.10.4    On-Site Measurements
   3.10.5    Postsampling Operations
   3.10.6    Calculations
   3.10.7    Maintenance
   3.10.8    Auditing Procedure
   3.10.9    Recommended Standards for
            Establishing Traceability
   3.10.10  Reference Method
   3.10.11   References
   3.10.12  Data Forms
13
5
3
3
18
7
2
1
1
5
1
6
1-04-82
1-04-82
1-04-82
1 -04-82
1-04-82
1-04-82
1-04-82
1-04-82
1-04-82
1-04-82
1-04-82
1-04-82
   3.11      Method 17—Determination of Paniculate
            Emissions from Stationary Sources
            (In-Stack Filtration Method)
   3.11.1    Procurement of Apparatus and Supplies
   3.11.2    Calibration of Apparatus
   3.11.3    Presampling Operations
   3.11.4    On-Site Measurements
   3.11.5    Postsampling Operations
   3.11.6    Calculations
   3.11.7    Maintenance
   3.11.8    Auditing Procedure
   3.11.9    Recommended Standards for
            Establishing Traceability
   3.11.10  Reference Method
   3.11.11  References
   3.11.12  Data Forms

   3.12     Method 9—Visible Determination of
            the Opacity  Emissions from
            Stationary Sources
9
2
3
6
1
1
2
2
1
11
1
1
-04-82
-04-82
-04-82
-04-82
-04-82
-04-82
-04-82
-04-82
-04-82
-04-82
-04-82
-04-82

-------
    June 1987
               Table of Contents (continued)

Section                                         Pages    Date
   3.12.1    Certification and Training of Observers         5        4-20-83
   3.12.2    Procurement of Apparatus and Supplies        2        4-20-83
   3.12.3    Preobservation Operations                    2        4-20-83
   3.12.4    On-Site Field Observations                   18        4-20-83
   3.12.5    Postobservation Operations                  2        4-20-83
   3.12.6    Calculations                                7        4-20-83
   3.12.7    Auditing Procedure                          2        4-20-83
   3.12.8    Reference Method                           5        4-20-83
   3.12.9    References and Bibliography                  1        4-20-83
   3.12.10  Data Forms                                 9        4-20-83

   3.13     Methods 6A and 6B—Determinations
            of Sulfur Dioxide, Moisture, and Carbon
            Dioxide Emissions from Fossil Fuel
            Combustion Sources

   3.13.1    Procurement of Apparatus and Supplies        18        7-01-86
   3.13.2    Calibration of Apparatus                     14        7-01-86
   3.13.3    Presampling Operations                     6        7-01-86
   3.13.4    On-Site Measurements                      25        7-01-86
   3.13.5    Postsampling Operations                     15        7-01-86
   3.13.6    Calculations                                9        7-01-86
   3.13.7    Maintenance                               3        7-01-86
   3.13.8    Auditing Procedure                          11        7-01-86
   3.13.9    Recommended Standards for                  1        7-01-86
            Establishing Traceability
   3.13.10  Reference Method                           5        7-01-86
   3.13.11   References                                 2        7-01-86
   3.13.12  Data Forms                                 18        7-01-86

   3.14     Method 7A—Determination of Nitrogen
            Oxide Emissions from Stationary
            Sources (Grab Sampling—Ion
            Chromatographic Method)

   3.14.1    Procurement of Apparatus and Supplies        10        7-01-86
   3.14.2    Calibration of Apparatus                     14        7-01-86
   3.14.3    Presampling Operations                     6        7-01-86
   3.14.4    On-Site Measurements                      7        7-01-86
   3.14.5    Postsampling Operations                     11        7-01-86
   3.14.6    Calculations                                6        7-01-86
   3.14.7    Maintenance                               2        7-01-86
   3.14.8    Auditing Procedure                          6        7-01-86
   3.14.9    Recommended Standards for                  1        7-01-86
            Establishing Traceability
   3.14.10  Reference Method                           3        7-01-86
   3.14.11   References                                 2        7-01-86
   3.14.12  Data Forms                                 12        7-01-86

   3.1 5     Method 7D—Determination of Nitrogen
            Oxide Emissions from Stationary
            Sources (Alkaline-Permanganate—Ion
            Chromatographic Method)

   3.15.1    Procurement of Apparatus and Supplies        18        7-01-86
   3.15.2    Calibration of Apparatus                     20        7-01-86
   3.15.3    Presampling Operations                     6        7-01-86
   3.15.4    On-Site Measurements                      10        7-01-86
   3.15.5    Postsampling Operations                     13        7-01-86

-------
                                                   June 1987
              Table of Contents (continued)

Section                                      Pages    Date
   3.15.6   Calculations                             5       7-01-86
   3.15.7   Maintenance                            3       7-01-86
   3.15.8   Auditing Procedure                        6       7-01-86
   3.15.9   Recommended Standards for                1       7-01-86
          Establishing Traceability
   3.15.10 Reference Method                        9       7-01-86
   3.15.11  References                              2       7-01-86
   3.15.12 Data Forms                             11       7-01-86

-------
                                                                       Section 3.0.4
                                                                       Rev. 6/9/87
                                                                       Page 1
          3.0.4.  PROCEDURE FOR NBS-TRACEABLE CERTIFICATION OF COMPRESSED
                  GAS WORKING STANDARDS USED FOR CALIBRATION AND
                  AUDIT OF CONTINUOUS SOURCE EMISSION MONITORS
                  (Revised TraceabiHty Protocol No. 1)
                                     CONTENTS

Subsection       Title

 3.0.4.0         General Information

 3.0.4.1         Procedure Gl;  Assay and Certification of
                 a Compressed Gas Standard Without Dilution

 3.0.4.2         References
                             4.0  GENERAL INFORMATION

4.0.1  Purpose and Scope of the Procedure

     Section 3.0.4 describes a procedure  for  assaying the concentration of gaseous
pollutant concentration standards and  certifying  that the assay concentrations are
traceable to an authoritative reference  concentration  standard.  This procedure is
recommended for certifying the local working concentration standards required by the
pollutant monitoring regulations of 40 CFR  Part 601'2 for the calibration and audit
of continuous source emission monitors.   The procedure covers certification of com-
pressed gas (cylinder) standards for CO, C02, NO, N02, and S02 (Procedure Gl).

4.0.2  Reference Standards

     Part 60 of the  monitoring  regulations1'2  require that working standards used
for calibration and audit  of  continuous  source  emission monitors be traceable to
either a National  Bureau  of  Standards  (NBS)  gaseous Standard Reference Material
(SRM) or a NBS/EPA-approved Certified  Reference  Material (CRM)3.  Accordingly, the
reference standard used for  assaying  and  certifying  a working standard for these
purposes must be an SRM, a  CRM,  or  a suitable intermediate standard (see the next
paragraph).  SRM cylinder gas standards  available  from NBS are listed in Table 7.2
at the end of subsection 4.0.  A  current list of CRM cylinder gases and CRM vendors
is available from the  Quality  Assurance Division (MD-77), Environmental Monitoring
Systems Laboratory, U. S. EPA, Research Triangle Park, NC 27711.
     The EPA regulations define  a  "traceable"  standard  as one which "...has been
compared and certified, either directly or  via not more than one intermediate stan-
dard, to a primary standard such  as a...NBS [gaseous] SRM or...CRM"4'5.  Certifica-
tion of a working standard directly to an SRM or CRM primary standard is, of course,
preferred and recommended because  of  the  lower  error.   However, an intermediate
reference standard is permitted, if necessary.    In particular,  a Gas Manufacturer's
Intermediate Standard (see subsection 4.0.2.1)   that has been referenced directly to
an SRM or a CRM according to Procedure Gl 1s an acceptable Intermediate standard and
could be used as the reference standard  on that basis.  However, purchasers of com-

-------
                                                                       Section 3.0.4
                                                                       Rev. 6/9/87
                                                                       Page 2


merdal gas standards referenced to an  Intermediate  standard such as a GMIS should
be aware that, according to the above  definition,  such a standard would have to be
used directly for calibration or audit.  Since a second Intermediate standard Is not
permitted, such a standard could  not  be  used  as  a reference standard to certify
other standards.

   4.0.2.1  Gas Manufacturer's Intermediate Standard (GMIS).  A GMIS 1s a compressed
   (cylinder) gas standard that has been assayedwith direct reference to an SRM or
   CRM and certified according to  Procedure  Gl,  and  also meets the following re-
   quirements:

     1.  A candidate GMIS must  be  assayed  a minimum of three (3) times, uniformly
     spaced over a three (3) month period.

     2.  Each of the three (or more)  assays  must be within 1.0 percent of the mean
     of the three (or more) assays.

     3.  The difference between the last  assay  and the first assay must not exceed
     1.5 percent of the mean of the three (or more) assays.

     4.  The GMIS must be  recertified  every  three months, and the reassay must be
     within 1.5 percent of the previous certified assay.  The recertified concentra-
     tion of the GMIS 1s the  mean  of  the previous certified concentration and the
     reassay concentration.

   4.0.2.2  Recert1f1cation  of  Reference  Standards.   Recert1f1cat1on requirements
   for SRMs and CRMs are specifiedByNBSand NBS/EPA,  respectively.  See 4.0.2.1
   for GMIS recertlflcation requirements.

4.0.3  Using the Procedure

     The assay/certification procedure described here 1s carefully designed to mini-
mize both systematic and random errors 1n  the assay process.  Therefore, the proce-
dure should be carried out as closely as possible to the way 1t 1s described.  Simi-
larly,  the assay apparatus  has  been  specifically  designed to minimize errors and
should be configured as closely as  possible  to the design specified.   Good labora-
tory practice should be observed 1n  the  selection of Inert materials (e.g. Teflon,
stainless steel, or glass,  1f  possible) and clean, non-contaminating components for
use 1n portions of the  apparatus  1n  contact  with  the candidate or reference gas
concentrations.

4.0.4  Certification Documentation

     Each assay/certification must be  documented  1n a written certification report
signed by the analyst and containing at least the following Information:

   1.  Identification number (cylinder number).

   2.  Certified concentration of the standard, 1n ppm or mole percent.

   3.  Balance gas 1n the standard mixture.

-------
                                                                       Section 3.0.4
                                                                       Rev. 6/9/87
                                                                       Page 3


   4.  Cylinder pressure at certification.

   5.  Date of the assay/certification.

   6.  Certification expiration date (see 4.0.6.3).

   7.  Identification of the reference  standard used:  SRM number, cylinder number,
   and concentration for an  SRM;  cylinder  number  and  concentration for a CRM or
   GMIS.

   8.  Statement that the  assay/certification  was performed according to this Sec-
   tion 3.0.4.

   9.  Identification of the  laboratory  where  the  standard was certified and the
   analyst who performed the certification.

   10. Identification of the gas analyzer  used for the certification, Including the
   make, model, serial number, the measurement  principle,  and the date of the last
   multipoint calibration.

   11. All analyzer readings  used  during  the assay/certification and the calcula-
   tions used to obtain the reported certified value.

   12. Chronological record of all certifications for the standard.

Certification concentrations should be reported to 3 significant digits.  Certifica-
tion documentation should be maintained for at least 3 years.

4.0.5  Certification Label

     A label or tag bearing the Information  described  1n Items 1 through 9 of sub-
section 4.0.4 must be attached to each certified gas cylinder.

4.0.6  Assay/Certification of Compressed Gas (Cylinder) Standards

   4.0.6.1  Aging of newly-prepared  gas  standards.   Freshly prepared gas standard
   concentrations and newly filled gascylindersmust be aged before being assayed
   and certified.  S02 concentrations contained  1n  steel cylinders must be aged at
   least 15 days; other standards must be aged at least 4 days.

   4.0.6.2  Stability test  for  reactive  gas  standards.   Reactive gas standards,
   Including nitric oxide (NO),nitrogendioxide(N02)• sulfur dioxide ($03),  and
   carbon monoxide (CO),  that have not  been previously certified must be tested for
   stability as follows:   Reassay the concentration  at least 7 days after the first
   assay and compare the two assays.  If the second assay differs from the first as-
   say by 1.5% or less,  the cylinder  may  be considered stable, and the mean of the
   two assays should be reported as the certified concentration.  Otherwise, age the
   cylinder for a week or more and  repeat  the test, using the second and third as-
   says as 1f they were the first and  second assays.  Cylinders that are not stable
   may not be sold and/or used for calibration or audit purposes.

-------
                                                                    Section 3.0.4"
                                                                    Rev. 6/9/87
                                                                    Page 4
4.0.6.3  Recert1f1cation of compressed  gas  standards.  Compressed gas standards
must be recertified according to this Section 3.0.4 within the time limits speci-
fied in Table 7.13«6«7.  The reassay  concentration must be within 5% of the pre-
vious certified concentration.  If not, the cylinder must be retested for stabil-
ity (subsection 4.0.6.2).  The  certified concentration of a recertified standard
should be reported as the mean of all assays, unless a clear trend or substantial
change suggests that previous assays are no longer valid.


       Table 7.1  Recertification limits for compressed gas standards.
                                                            Maximum months until
                                                            recertification for
Pollutant
Carbon monoxide
Nitric oxide
Sulfur dioxide
Nitrogen dioxide
Carbon dioxide
Oxygen
Sulfur dioxide and
carbon dioxide
Propane
Others not specifically
Balance
gas
N2 or air
N2
N2
\\2 or air
N2 or air
N2
N2
N2 or air
listed
Concentration
range
I 5 ppm
I 10 ppm
I 10 ppm
£ 10 ppm
I 300 ppm
I 2 percent
* 200 ppm S02,
* 10 percent C02
£ 5 ppm

cylinder material :
Al or SS other
18
18
18
6
18
18
18
18
6
6
6
6
6
18
18
6
6
6
4.0.6.4  Minimum cylinder pressure.  No compressed  gas cylinder  standard should
be used when its gas pressureis~ below  700  kPa (100 psi), as indicated by the
cylinder pressure gauge.

4.0.6.5  Assay/certification of multi-component compressed gas standards.  Proce-
dure Gl may be used to assay and certify individual  components of multi-component
gas standards, provided that  none  of  the  components  other than the component
being assayed cause a detectable response on the analyzer.

-------
                                                                       Section 3.0.4
                                                                       Rev.  6/9/87
                                                                       Page  5
4.0.7  Analyzer Calibration
   4.0.7.1  Basic analyzer calibration  requirements.  The assay procedure described
   in this Section 3.0.4 employs a direct ratio referencing technique that inherent-
   ly corrects for minor analyzer calibration variations (drift) and DOES NOT depend
   on the absolute accuracy of the  analyzer  calibration.   What is required of the
   analyzer is as follows:  1) it  must  have  a linear response to the pollutant of
   interest (see subsection 4.0.7.5), 2) it must have good resolution and low noise,
   3) its response calibration must be reasonably stable during the assay/certifica-
   tion process, and 4) all  assay  concentration  measurements must fall within the
   calibrated response range of the analyzer.

   4.0.7.2  Analyzer multipoint calibration.   The  gas analyzer used for the assay/
   certification must have had a multipoint  calibration  within 3 months of Its use
   when used  with this  procedure.  This calibration is not used to  quantitatively
   interpret analyzer readings during  the  assay/certification of the candidate gas
   because a more accurate, direct  ratio  comparison of the candidate concentration
   to the reference standard concentration is  used.  However, this multipoint cali-
   bration is necessary to establish  the  calibrated  range of the analyzer and its
   response linearity.
        The multipoint calibration should consist  of analyzer responses to at least
   5 concentrations, including zero, approximately evenly spaced over the concentra-
   tion range.  Analyzer response units  may be volts, millivolts, percent of scale,
   or other measurable analyzer response units.   The upper range limit of the cali-
   brated range is determined by the highest calibration point used.  If the analyz-
   er has a choice  of  concentration  ranges,  the  optimum range for the procedure
   should be selected and calibrated.   Plot  the calibration points and compute the
   linear regression slope and intercept.   See subsection 4.0.7.5 for linearity re-
   quirements and the use of a  mathematical  transformation, 1f needed.  The Inter-
   cept should be less than 1  percent  of  the upper concentration range limit, and
   the correlation coefficient (r) should be at least 0.999.

   4.0.7.3  Zero and span check and adjustment.   On each day that the analyzer will
   be used for assay/certification, its response  calibration must be checked with a
   zero and at least one span  concentration near the upper concentration range Urn-
   it.  If necessary, the zero and  span controls of the analyzer should be adjusted
   so that the  analyzer's  response  (i.e.  calibration  slope)  1s within about ±5
   percent of the response indicated by  the most recent multipoint calibration.  If
   a zero or span adjustment is made,  allow  the analyzer to stabilize for at least
   an hour or more before beginning  the assay procedure, since some analyzers drift
   for a period of time following zero  or  span adjustment.  If the analyzer is not
   in continuous operation, turn it on  and  allow  1t  to stabilize for at least 12
   hours before the zero and span check.

   4.0.7.4  Pollutant standard for multipoint  calibration and zero and span adjust-
   ment.  The pollutantstandardorstandardsused for multipoint calibration or
   zero and span checks or adjustments must  be obtained from a compressed gas stan-
   dard certified traceable to an NBS SRM or a NBS/EPA CRM according to Procedure Gl
   of this Section 3.0.4.  This standard need not be the same as the reference stan-
   dard used in the assay/certification.  The zero gas must meet the requirements 1n
   subsection 4.0.8.

-------
                                                                       Section 3.0.'
                                                                       Rev. 6/9/87
                                                                       Page 6


   4.0.7.5  Linearity of analyzer response.  The direct  ratio assay  technique used
   In Procedure 61 requires tnat the  analyzer  have a linear response to concentra-
   tion.  Linearity 1s determined by comparing the quantitative difference between a
   smoothly-drawn calibration curve based on  all  calibration points and a straight
   line drawn between zero and an upper  reference  point (see Figure 1).  This dif-
   ference 1s measured in concentration  units,  parallel to the concentration axis,
   from a point on the  calibration  curve  to  the corresponding point for the same
   response on the straight line.
        For the general linearity  requirement,  the  straight line is drawn between
   zero and the highest calibration point (Figure la).  Linearity is then acceptable
   when no point on the smooth calibration  curve deviates from the straight line by
   more than 1.5 percent of the value  of the highest calibration concentration.  An
   alternative linearity requirement is defined on the basis of the actual reference
   and candidate concentrations to be used for  the assay.  In this case, the refer-
   ence and candidate concentrations are  plotted  on the calibration curve, and the
   straight line is drawn from zero to the reference concentration and extrapolated,
   1f necessary, beyond the candidate  concentration  (Figure Ib).  The deviation of
   the smooth calibration curve from  the  straight line at the candidate concentra-
   tion point then must not exceed 0.8 percent of the value of the reference concen-
   tration.  This latter  specification  may  allow  the  use  of an analyzer having
   greater nonlinearlty when the  reference  and candidate concentrations are nearly
   the same.
        For analyzers having  an  inherently  non-linear  response, the response can
   usually be linearized with a  simple  mathematical transformation of the response
   values, such as R'= square  root(R)  or  R'=  log(R), where R' is the transformed
   response value and R 1s  the  actual  analyzer  response value.  Using the trans-
   formed response values, the multipoint  calibration  should meet one of the above
   linearity requirements as well as  the requirements for intercept and correlation
   coefficient given 1n subsection 4.0.7.2.

4.0.8  Zero Gas

     Zero gas used for dilution  of  any  candidate  or reference standard should be
clean, dry, zero-grade air or  nitrogen  containing a concentration of the pollutant
of Interest equivalent to less than 0.5  percent of the analyzer's upper range limit
concentration.  The zero gas also  should  contain  no contaminant that causes a de-
tectable response on the  analyzer  or  that  suppresses  or enhances the analyzer's
response to the pollutant.  The  oxygen  content  of  zero air should be the same as
that of ambient air.

4.0.9  Accuracy Assessment of Commercially Available Standards

     Periodically, the USEPA  will   assess  the  accuracy  of commercially available
compressed gas standards that  have  been  assayed  and  certified according to this
Section 3.0.4.  Accuracy will  be  assessed  by EPA audit analysis of representative
actual commercial  standards obtained  via  an  anonymous  agent.  The accuracy audit
results, Identifying the actual gas  manufacturers  or vendors,  will be published as
public Information.

-------
                                                                                                            Section 3.0.4
                                                                                                            Rev.  6/9/87
                                                                                                            Page  7
                                o
                                CL
                               _>
                                to
                                c
                                i
                              Concentration difference
                              between calibration curve
                              and straight line must
                              not exceed 1.5% of
                              Cmax at any point
                              between 0 and
                          —  Crnax
                            Concentration
                                points
                                                                       Smooth calibration
                                                                       curve based on all
                                                                       calibration
                                                                       points
                                                                             V
                                                                                   i
                                                            Straight line
                                                            between zero
                                                            and highest
                                                            calibration
                                                            point
                                                                             Concentration
                                                                             at highest
                                                                               libration
                                                                           f point (Cmax)
                                              Concentration, ppm or percent
                                               a) General linearity requirement
a
5
ce
        Concentration difference
        between calibration
        curve and straight line
        at candidate
        concentration point   \.
        must not exceed 0.8%    J^
        of the reference
        concentration
                  Candidate
              concentration
                      Smooth calibration
                      curve based on all
                      calibration points
Calibration
  points
                         Straight line
                         between zero
                         and reference
                         concentration
                                       Reference
                                       concentration
                                                                   <
Concentration difference
between calibration
curve and straight line
at candidate concentration
point must not       v.
exceed 0.8% of the      >^
reference concentration
  Smooth calibration
  curve based on all
  calibration points
                                                                                                  Candidate
                                                                                                  concentration
  Calibration
     points
Reference
concentration
                                                                                             Straight line between zero and
                                                                                             reference concentration, extrapolated
                                                                                             to beyond the candidate concentration
                   Concentration, ppm or percent
                                                                                      Concentration, ppm or percent
                                             b) Alternative linearity requirement
                          Figure  1.    Illustration of  linearity  requirements.

-------
                       Table 7.2.  NBS SRM reference gases.
                                           Section  3.0.4
                                           Rev.  6/9/87
                                           Page  8
SRM
number
2627
2628
2629
1683b
1684b
1685b
1686b
1687b
2630
2631

2653
2654
2655
2656

2612a
2613a
2614a

1677C
2635
1678c
1679c
2636
1680C
1681c
2637
2638
2639
2640
2641
2642

2657
2658
2659



NBS-SRM
Type
NO/N2
NO/N2
NO/N2
NO/N2
NO/N2
NO/N2
NO/N2
NO/N2
NO/N2
NO/N2

N02/Air
N02/Air
N02/Air
N02/A1r

C0/A1r
C0/A1r
CO/Air

CO/N2
CO/N2
CO/N2
CO/N2
CO/N2
CO/N2
CO/N2
CO/N2
CO/N2
CO/N2
CO/N2
CO/N2
CO/N2

02/N2
02/N2
02/N2



cylinders contain
Nominal
concentration
5
10
20
50
100
250
500
1000
1500
3000

250
500
1000
2500

10
20
45

10
25
50
100
250
500
1000
2500
5000
1
2
4
8

2
10
21



ppm
ppm
ppm
ppm
ppm
ppm
ppm
ppm
ppm
ppm

ppm
ppm
ppm
ppm

ppm
ppm
ppm

ppm
ppm
ppm
ppm
ppm
ppm
ppm
ppm
ppm
percent
percent
percent
percent

percent
percent
percent



SRM
number
1693
1694
1661a
1662a
1663a
1664a
1696

1670
1671
1672

2632
2633
2634
2619a
2620a
2621a
2622a
2623a
2624a
2625a
2626a
1674b
1675b

1665b
1666b
1667b
1668b
1669b

2643
2644
2645
2646
2647
2648
2649
2650
approximately 870 liters of
Type
S02/N2
S02/N2
S02/N2
S02/N2
S02/N2
S02/N2
S02/N2

C02/Air
C02/Air
C02/Ai r

C02/N2
C02/N2
C02/N2
C02/N2
C02/N2
C02/N2
C02/N2
C02/N2
C02/N2
C02/N2
C02/N2
C02/N2
C02/N2

C3H8/Air
C3H8/Air
C3H8/Air
C3H8/Air
C3H8/Air

C3H8/N2
C3H8/N2 .
C3H8/N2
C3H8/N2
C3H8/N2
C3H8/N2
C3H8/N2
C3H8/N2
gas at STP.
Nominal
concentration
50 ppm
100 ppm
500 ppm
1000 ppm
1500 ppm
2500 ppm
3500 ppm

330 . ppm
340 ppm
350 ppm

300 ppm
400 ppm
800 ppm
0.5 percent
1.0 percent
1.5 percent
2.0 percent
2.5 percent
3.0 percent
3.5 percent
4.0 percent
7.0 percent
14.0 percent

3 ppm
10 ppm
50 ppm
100 ppm
500 ppm

100 ppm
250 ppm
500 ppm
1000 ppm
2500 ppm
5000 ppm
1 percent
2 percent

For availability, contact:
Office of Standard Reference Materials
Chemistry Building, Room B311
NBS, Gaithersburg, Maryland  20899
(301) 975-6776.  (FTS 879-6776)

-------
                                                                       Section 3.0.4
                                                                       Rev. 6/9/87
                                                                       Page Gl-1


            4.1  PROCEDURE Gl;  ASSAY AND CERTIFICATION OF A COMPRESSED
                 GAS STANDARD WITHOUT DILUTION


4.1.1   Applicability

     This procedure may be used to assay the concentration of a candidate compressed
gas (cylinder) pollutant standard, based  on  the  concentration of a compressed gas
(cylinder) reference standard of the  same  pollutant compound, and certify that the
assayed concentration thus established  for  the  candidate standard 1s traceable to
the reference standard.  The procedure  employs  a pollutant gas analyzer to compare
the candidate and reference gas  concentrations by direct measurement—without dilu-
tion of either gas--to minimize assay error.

4.1.2   Limitations

   1.  The concentration of the candidate  gas  standard must be between 0.3 and 1.3
   times the concentration of the reference gas standard.

   2.  The analyzer must have a  calibrated range capable of directly measuring both
   the candidate and the reference gas concentrations.

   3. The analyzer's response (or transformed  response) must be linear with respect
   to concentration.

   4. The balance gas in both the  candidate and reference standards must be Identi-
   cal, unless it can be shown that the analyzer is Insensitive to any difference 1n
   the balance gases.

   5. A source of clean, dry zero gas is required.

4.1.3   Assay Apparatus

     Figure Gl illustrates the relatively simple assay apparatus.  The configuration
is designed to allow convenient routing of the zero gas and undiluted samples of the
reference gas and candidate  gases,  in  turn,  to  the analyzer for measurement, as
selected by three-way valves VI and  V2.   Pressure regulators and needle valves (V3
and V4) control the individual gas  flows.   The pollutant concentrations are deliv-
ered to the analyzer via a vented tee, which discharges excess flow and Insures that
the assay concentrations sampled by the analyzer are always at a fixed (atmospheric)
pressure.  A small, uncalibrated rotameter monitors the vent flow to verify that the
total  gas flow rate exceeds the sample  flow  rate demand of the analyzer so that no
room air is admitted through the vent.  Valves VI and V2 could be replaced by a sin-
gle four-way valve (with 3 inputs  and  1  output)  or by manually moving the output
connection to each of the gases as needed.  See also subsection 4.0.3.

4.1.4   Analyzer

     See subsection 4.0.7.1.  The pollutant gas analyzer must have a linear response
function and a calibrated range capable  of measuring the full  concentration of both
the candidate and the reference gas  standards  directly, without dilution.  It must

-------
                                                                Section  3:0.4
                                                                Rev. 6/9/87
                                                                Page Gl-2
                                              Vent
 Zero Gas
(Air or N2)
  Supply
                                               1
                                                  Rotameter
                                                        To Analyzer
                                         Three-Way Valve
                             Pressure
                            Regulators
               Reference
                 Gas
Candidate
   Gas
    Figure Gl.   Suggested assay apparatus for Procedure 61,

-------
                                                                       Section 3.0.4
                                                                       Rev. 6/9/87
                                                                       Page Gl-3


have good resolution  (readability), good precision,  a stable response, and low out-
put signal noise.  In  addition,  the  analyzer  must  have good specificity for the
pollutant of Interest so that 1t has  no detectable response to any contaminant that
may be contained 1n either the  candidate  or  reference  gas.  If the candidate and
reference gases contain dissimilar balance  gases  (air versus nitrogen or different
proportions of oxygen 1n the balance air,  for example), the analyzer must be proven
to be Insensitive to the two different  balance  gases.  This may be accomplished by
showing no difference 1n  analyzer  response when measuring pollutant concentrations
diluted with Identical flow rates of the two balance gases.
     The analyzer should be  connected  to  a  suitable, precision chart recorder or
other data acquisition device to  facilitate graphical observation and documentation
of the analyzer responses obtained during the assay.

4.1.5   Analyzer Calibration

   4.1.5.1  Multipoint calibration.  See subsections 4.0.7.2 and 4.0.7.4.

   4.1.5.2  Calibration range.   The  calibrated  range of the analyzer must Include
   both the candidate and reference gas concentrations, such that the higher concen-
   tration does not exceed 97 percent of  the  upper range limit, and the lower con-
   centration 1s not below 25  percent  of  the  upper range limit (assuming a lower
   range limit of zero).   Within  these  limits, select a calibrated analyzer range
   that will produce the highest analyzer responses.

   4.1.5.3  Linearity.   The  direct  ratio  assay  technique used In this procedure
   requires that the analyzer have  a  linear response to concentration (see subsec-
   tion 4.0.7.5).  High-concentration-range analyzers  of the type that are required
   for this procedure may not be Inherently  linear, but they usually have a predic-
   table, non-11 near response characteristic  that can be mathematically transformed
   to produce a sufficiently linear response characteristic suitable for use 1n this
   procedure.  Any such response transformation  should  be verified by using 1t for
   the multipoint calibration.  Caution  should  be exercised In using a transformed
   response curve because physical  zero  or  span  adjustments  to the analyzer may
   produce unexpected effects on the transformed characteristic.

   4.1.5.4  Zero and span adjustment.  See subsections  4.0.7.3 and  4.0.7.4.  Prior
   to carrying out the assay/certification  procedure,  check the calibration of the
   analyzer and, 1f necessary, adjust the  analyzer's  zero and span controls to re-
   establish the response characteristic  determined  at  the most recent multipoint
   calibration.  Allow the analyzer to stabilize  for an hour or more after any zero
   or span adjustment.  If there 1s any doubt that a transformed response character-
   istic 1s still  linear following a  zero or span adjustment,  verify linearity with
   a multipoint calibration (subsection  4.0.7.2)  using  at least 3 known pollutant
   concentrations,  Including zero.

4.1.6   Assay Gases

   4.1.6.1  Candidate gas standard.   See subsections 4.0.6 and  4.1.2.

   4.1.6.2  Reference gas standard.   See subsections 4.0.2,  4.1.2,  and 4.0.6.4.   Se-
   lect a reference standard such that the concentration of the candidate gas Is not

-------
                                                                       Section 3.0.4
                                                                       Rev. 6/9/87
                                                                       Page Gl-4


   more than 30 percent above nor  less  than  70 percent below the concentration of
   the standard.

   4.1.6.3  Zero gas.  See subsection 4.0.8.   The zero gas should match the balance
   gas used 1n the cylinder concentrations.

4.1.7   Assay Procedure

   1.  Verify that the assay apparatus  1s properly configured,  as described in sub-
   section 4.1.3 and shown 1n Figure Gl.

   2.  Verify that the linearity of the  analyzer has been checked within the last 3
   months (see subsections 4.0.7.2, 4.0.7.5, and  4.1.4),  that the zero and span are
   adjusted correctly (subsection  4.0.7.3),  that  the  candidate and reference gas
   concentrations are within 25 and 97 percent of the upper range limit of the cali-
   brated measurement range of  the  analyzer,  and  that  the analyzer 1s operating
   stably.

   3.  Adjust the flow rates of  the three gases (reference,  candidate, and zero) to
   approximately the same value that will  provide  enough flow for the analyzer and
   sufficient excess to assure that no ambient air will be drawn Into the vent.

   4.  Conduct a triad of measurements with  the analyzer.  Each triad consists of a
   measurement of the zero  gas  concentration,  a  measurement of the reference gas
   concentration, and a measurement of the  candidate gas concentration.  Use valves
   VI and V2 to select each of  the  three concentrations for measurement.  For each
   measurement, allow ample time for the  analyzer to achieve a stable response read--
   Ing.  Record the stable  analyzer  response  for each measurement, using the same
   response units (volt, millivolts, percent of scale,  etc.)  used for the multipoint
   calibration and any transformation of  the response readings necessary for linear-
   ity.  Do not  translate  the  response  readings  to concentration values via the
   calibration curve (see the  footnote  following  Equation  Gl).   Do not make any
   zero,  span, or other physical  adjustments  to  the  analyzer during the triad of
   measurements.

   5.  Conduct at least 2  additional  measurement  triads,  similar to step 4 above.
   However,  for these subsequent triads,  change  the order of the three measurements
   (e.g.  measure reference gas, zero  gas,  candidate  gas  for the second triad and
   zero gas, candidate gas, reference gas for the third triad,  etc.).

   6.  If any one or more of the  measurements of a triad is Invalid or abnormal  for
   any reason, discard all three measurements of the triad and repeat the triad.

   7.  For each  triad  of  measurements,  calculate  the assay  concentration of the
   candidate gas as follows:


                                  Cc = cr RC " RZ                        Equation Gl

-------
                                                                       Section 3.0.4
                                                                       Rev. 6/9/87
                                                                       Page Gl-5
   where:  Cc = Assay concentration of the candidate gas standard, ppm or
                  percent;
           Cr = Concentration of the reference gas standard,  ppm or percent;
           Rc = Stable response reading of the analyzer for the candidate
                  gas, analyzer response units;*
           Rz = Stable response reading of the analyzer for the zero gas,
                  analyzer response units;*
           Rr = Stable response reading of the analyzer for the reference
                  gas, analyzer response units.*

      *Analyzer response units are  the  units  used  to express the direct response
       readings of the analyzer, such  as  volts,  millivolts,  percent of scale,  etc.
       DO NOT convert these direct response readings to concentration units with the
       multipoint calibration curve or  otherwise   adjust  these readings except for
       transformation necessary to achieve response linearity.

   8.  Calculate the mean of the  3  (or  more) valid assays.   Calculate the percent
   difference of each  assay from the mean.  If any one of the  assay values differs
   from the mean by more  than  1.5%,  discard  that assay value and conduct another
   triad of measurements to obtain another assay  value.  When at least 3 assay  val-
   ues all agree within 1.5% of their  mean,   report the mean  value as the certified
   concentration of the candidate gas  standard.   For newly-prepared reactive stan-
   dards, a reassay at least 7 days later  is required to check the stability of the
   standard; see subsection 4.0.6.2.

4.1.8  Stability Test for Newly-Prepared Standards

     See subsections 4.0.6.1 and 4.0.6.2.

4.1.9  Certification Documentation

     See subsections 4.0.4 and 4.0.5.

4.1.10 Recertification Requirements

     See subsections 4.0.6.3 and 4.0.6.4.

-------
                                                                      Section 3.0.4
                                                                      Rev. 6/9/87
                                                                      References
4.2    References.
   1.  Code of Federal Regulations, Title 40, Part 60, "Standards of Performance for
   New Stationary Sources," Appendix A, Method 20 (1982).

   2.  Standards of Performance  for  New  Stationary Sources; Quality Assurance Re-
   quirements for Gaseous Continuous Emission Monitoring Systems Used for Compliance
   Determination, promulgated in  the  Federal  Register,  June  4, 1987, pp. 21003-
   21010.

   3.  "A Procedure for Establishing Traceability of Gas Mixtures to Certain Nation-
   al Bureau of Standards  Standard  Reference  Materials.  EPA-600/7-81-010.  Joint
   publication by NBS and  EPA,  May  1981.    Available from the U.S. Environmental
   Protection Agency, Environmental Monitoring  Systems Laboratory (MD-77), Research
   Triangle Park, NC  27711.

   4.  Code of Federal Regulations, Title 40, Part 50, "National Ambient Air Quality
   Measurement Methodology".

   5.  Code of Federal Regulations, Title 40, Part 58, "Ambient Air Quality Surveil-
   lance," Appendixes A and B.

   6.  Shores, R. C. and  F.  Smith, "Stability Evaluation of Sulfur Dioxide, Nitric
   Oxide, and Carbon Monoxide Gases in Cylinders.  NTIS No. PB 85-122646.  Available
   from the National Technical  Information  Service,  5285 Port Royal Road, Spring-
   field, VA  22161.

   7.  Method 6A  and  6B,  "Determination  of  Sulfur Dioxide, Moisture, and Carbon
   Dioxide Emissions from Fossil  Fuel  Combustion Sources," Quality Assurance Hand-
   book for Air  Pollution  Measurement  Systems.  Volume  III, Section 3.13.8,  July
   1986.Available from the U.S.Environmental  Protection Agency, Center for Envi-
   ronmental  Research Information,  Cincinnati, OH  45268.

   8.  "List  of  Designated  Reference  and  Equivalent  Methods."  Current edition
   available from the U.S. Environmental  Protection Agency, Environmental Monitoring
   Systems Laboratory, Quality Assurance  Division  (MD-77), Research Triangle Park,
   NC  27711.

-------
                                            Section No. 3.0.7
                                            Date November 5,  1985
                                            Page 1

7.0  CALCULATION AND INTERPRETATION OF ACCURACY FOR CONTINUOUS
     EMISSION MONITORING SYSTEM (CEMS)

    This section contains a discussion  on  the accuracy calcula-
tions required  in  Appendix  F   and  their interpretation.   The
goals  of Appendix F, Procedure 1, are to (1) assess CEMS  accur-
acy, (2) indicate when a CEMS is out-of-control and correction is
required,  and (3) specify criteria for unacceptable  CEMS  data.
The quarterly accuracy assessments required in Appendix F provide
a  mechanism  for  identifying  and  correcting  CEMS's that  are
out-of-control.   This  results in an increase in acceptable CEMS
data.  Increasing acceptable CEMS data strengthens decisions made
with regard to compliance.

    The following  subsections  discuss  the meaning, interpreta-
tion, calculation, and reporting of accuracy data.

7.1  Meaning of Accuracy

    Accuracy is  the measure of the closeness of a measurement to
its "true value."  Although the true value is not known,  it  can
be approximated by the use of  an  appropriate standard of refer-
ence,  for  example,  an NBS-SRM (National Bureau of Standards
Standard  Reference  Materials),  a  primary standard.  Secondary
standards are also used as an  approximation to "truth," although
errors may be introduced in this process.

    The  preferred  measure of accuracy depends on the situation.
If the magnitude of  the  difference tends to be dependent on the
true value, T, then the percentage  difference is preferable.  If
it is desired to follow or observe the pattern of the differences
over time,  then  the  signed  difference  or  signed  percentage
difference is preferable.

    In the context of accuracy  data  based  on Appendix F, three
types of audits  for  CEMS  accuracy  assessment  are  specified:
Relative Accuracy  Test  Audits  (RATA), Relative Accuracy Audits
(RAA), and Cylinder Gas Audits (CGA).  The procedure for the RATA
and  the  RAA  are  the  same as for the Relative  Accuracy  Test
described in the applicable EPA performance  specification (e.g.,
Performance Specification  2  for  S02  and  NO , and Performance
Specification 3 for 02 and CO2), with the exception  that the RAA
requires  three  rather than nine sets of measurements,  and  the
accuracy  is  based on the average of the three sets of data.  In
addition,   EPA   performance  audit  samples  must  be  analyzed
concurrently  with  the  RATA samples to demonstrate and document
the proficiency and accuracy  of the analytical system.  The same
person  must  conduct the RATA and the EPA audit sample analysis.
Thus, the RATA approximates  "truth" by the reference method test
results, which are in turn checked for analytical accuracy by EPA
audit sample analyses.  The EPA audit  sample analysis must agree

-------
                                            Section No. 3.0.7
                                            Date November 5, 1985
                                            Page 2

within 5 percent of the  audit  concentration  on each of two S09
audit samples or within 10 percent of the  audit concentration on
each of two NO  audit samples.

    In  Appendix F, each GEMS must be audited at least once  each
calendar quarter.  Successive audits  shall  occur no closer than
two months apart.  The audits must be conducted as follows:
    1.
The RATA  must  be
calendar quarters.
the   Performance
 conducted  at  least  once  every four
 The  RATA is conducted as described in
Specifications  in  Appendix  B  (e.g.,
                                                               In
Performance   Specification   2  for  S02  and  NO ).
addition,  the  appropriate  performance   audit  xsamples
received  from  EPA  are  analyzed  as  described  in  the
applicable  Reference Methods (e.g., Methods 6 for S00 and
7 for NO ).                                          z
        *£

If applicable, a CGA may be conducted in three of the four
calendar quarters.  A CGA is conducted by  challenging the
GEMS's  (both pollutant and diluent  monitors,  if  appli-
cable) with  an  audit  gas  of known concentration at two
points within the following ranges:
Audit
point
1
2
Audit range
Pollutant monitors
20 to 30% of span value
50 to 60% of span value
Diluent monitors for--
co2 o2
5 to 8% by
volume
10 to 14% by
volume
4 to 6% by
volume
8 to 12% by
volume
        A  separate  audit  gas cylinder must be used  for  audit
        points  1  and  2.  No dilution of the gas from the audit
        cylinder is allowed when challenging the GEMS.  Challenge
        the GEMS three times at each point,   and  use the average
        of  the three responses  in  determining  accuracy.   The
        monitor   should  be  challenged  at  each  point  for  a
        sufficient   period   of   time   to  assure  absorption-
        desorption  of  the  CEMS  sample  transport surfaces has
        stabilized.   Each  monitor  is  audited  in  its  normal
        sampling mode, i.e.,  pass  the  audit  gas  through  all
        filters,  scrubbers,  conditioners,    and  other  monitor
        components used during normal sampling and as much of the
        sampling probe as is practical.  At a minimum,  the audit
        gas should be introduced  at  the  connection between the
        probe and the sample line.  Audit gases must be certified

-------
                                            Section No.  3.0.7
                                            Date November 5,  1985
                                            Page 3

        by  comparison  with  gaseous NBS-SRM or NBS/EPA approved
        CRM   (Certified   Reference   Material)   following  EPA
        Traceability  Protocol No.  1.  Procedures for preparation
        of  CRM's  are described in Reference 2.  Procedures  for
        preparation of EPA Traceability Protocol  No. 1  gases are
        described  in  Reference  3.   The difference between the
        actual concentration of the audit  gas  and  the  concen-
        tration  indicated  by  the monitor is used to assess the
        accuracy of the CEMS.

    3.  The RAA may  be  conducted  three  of  the  four calendar
        quarters.   To  conduct  a  RAA,  follow  the  procedures
        described in the applicable Performance Specification  in
        Appendix B for the Relative  Accuracy  Test,  except that
        only   three  sets  of  measurement  data  are  required.
        Analysis of EPA performance audit samples is required for
        the RAA.  The relative difference between the mean of the
        reference method values and the mean  of  the CEMS values
        (in  terms  of  the standard)  are  used  to  assess  the
        accuracy of the CEMS.

    The  performance  of RATA's, RAA's,  and  CGA's  provides  an
independent check of the CEMS accuracy.  These independent audits
serve to document  that  the  CEMS  is  providing  quality  data.
Examples of audit  calculations  are given in the subsection that
follows.

    In  summary, an accuracy  assessment  is  a  measure  of  the
deviation of a measurement  obtained  under  standard operational
procedures from a  known  reference  measurement.   There  is  no
reason  to  expect that accuracy will remain constant  over  each
quarter because of changes in  calibration  gases,  analysts, and
environment.

7.2  Example Calculations and Interpretation for Accuracy

7.2.1  Relative Accuracy  Test  Audit Calculations - Example data
from a RATA on a S02/02 CEMS are shown in Table 7.1.

    The SO2 and 02 CEMS data shown in Table 7.1'were corrected to
a dry basis using Equation 7-1:

                        CEMS_                       Equation 7-1
         CEMS     ^   =	ppm, wer
             ppm, dry       _
                               ws
where
         B   = moisture fraction of the CEMS gas sampled.

-------
TABLE 7.1
                                            Section No. 3.0.7
                                            Date November 5, 1985
                                            Page 4
     RELATIVE ACCURACY  TEST AUDIT DATA  FOR  S02 AND  C>2  CEMS

Run

number
1
2
3
4
5
6
7
8
9
Avg
S02
RM

ppm
500
505
510
510
500
500
510
505
510
	
S02
CEMS , ,

ppm
475
480
480
480
480
500
510
505
520
	
°2
RM

%Q
3.0
3.0
3.0
2.9
2.9
3.0
3.0
2.9
2.9
	
°2
CEMS.,


3.1
3.1
3.0
2.9
3.0
3.1
3.1
3.0
3.0
	
S°2
RM,,

ng/J
422.4
426.6
430.8
428.4
420.0
422.4
430.8
424.2
428.4
426.0
so2
CEMS ,,

ng/Ja
403.5
407.7
405.4
403.2
405.4
424.7
433.3
426.6
439.3
413.1
so2
Diff,

ng/J
18.9
18.9
25.4
25.2
14.6
-2.3
-2.5
-2.4
-10.9
9.43
    RM
,  =  reference  method  data,  dry  basis,

,  =  monitor  data,  dry basis.
  CEMS

  The  S02  and 0« CEMS and RATA data in Table 7.1 were converted
to the units of tne applicable standard using Equation 7-2:
         E = CF
                       20.9
                 20.9 - percent
                                              Equation  7-2
where
Percent 0,
  E  = pollutant  emission, ng/J  (Ib/million  Btu),
                                      3
  C  = pollutant  concentration,  ng/dsm  (Ib/dscf),

  F  = factor  representing a  ratio  of  the  volume  of  dry
      flue  gas generated to  the calorific  value of the
      fuel, dsm  /J  (dscf/million Btu),  and

      oxygen  content  by volume  (expressed as  percent),
      dry basis.
  Note;  For  the calculations shown in Table 7.1, ppm of S02 wa^
converted  to  ng/J  using  a  conversion  fagtor of 2.66  x  10
ng/scm/ppm and an F factor of 2.72 x 10   dsm /J.

-------
                                            Section No.  3.0.7
                                            Date  November  5, 1985
                                            Page  5

  For complete explanation of the equations  and calculations,  see
40  CFR; Part 60; Appendix  A;  Method  19;   5.   Calculation  of
Particulate, Sulfur Dioxide, and Nitrogen Oxides  Emission  Rates.

  After the  data are converted to the units of the standard,  the
Relative  Accuracy  (RA)  is calculated by using  the equations in
Section  8  of  Performance  Specification 2.   For  convenience in
illustrating the calculation,  these equations (7-3   through  7-8)
are also shown here.

  The  average  difference, d, is calculated for  the S02  monitor
using Equation 7-3:
         _                   -    n
         d = ~  2 (Xi " Yi)       2  di              Equation 7-3
             n .  ,            n  . ,


           = - (84.9) = 9.43 ng/J
             9

where

         n  = number of data points,

         X. = concentration from reference method (RM, in Table
          1   7.1),  ng/J,                             a

         Y. = concentration from the CEMS (CEMS, in Table 7.1),
          1   ng/J,                             d

         d. = signed difference between individual pairs,  X. and
              Y±, ng/J, and

        Zd. = algebraic sum of the individual differences,  d.,
              ng/J.                                          1


The standard deviation S. is calculated using Equation 7-4:
                  Vi rn   2   i    n     2!
                      S di       (  2  di>            Equation 7-4
                 n"1U-i      n   i-i     J

-------
             V
    '1         1        21
     -  2344 - -  (84.9)
     8         9         I
   Section No. 3.0.7
   Date November 5, 1985
   Page 6

= 13.9 ng/J.
    The 2.5 percent error confidence coefficient, CC, is calcula-
ted using Equation 7-5:
         CC = t
               0.975
                                         Equation 7-5
= 2.306
                          = 10.68 ng/J.
where tQ g75 = t-values in Table 7.2 for n = 9.
         TABLE 7.2.  VALUES OF t FOR 95 PERCENT PROBABILITY0
                                        .Area = 0.95
\rea = 0

na
2
3
4
5
6
.025^

to.
12.
4.
3.
2.
2.
"^.^
-rtfirrffflll
^0.975
975
706
303
182
776
571
*^"^
^"""^x. .^Area = 0.025
ItmT
0
a
n t0.975
7 2.447
8 2.365
9 2.306
10 2.262
11 2.228
t0.975
na
12
13
14
15
16

*0
2
2
2
2
2

.975
.201
.179
.160
.145
.131
           The values in this table are already corrected  for
           n-1 degrees of freedom.  Use n equal to the number of
           individual values.

-------
                                            Section No. 3.0.7
                                            Date November 5, 1985
                                            Page 7

       The RA for the RATA is calculated using Equation 7-6:
         RA =
                        Iccl
             x 100
Equation 7-6
                    RM
|9.43|
                         . 68|
                               x 100 = 4.72%
                    426
where
         RA = relative accuracy, %,

          ld|= absolute value of the mean differences .from Equa-
              tion 7-3, ng/J,

         |CC|= absolute value of the confidence coefficient  from
              Equation 7-5, ng/J, and

         KM = average reference method value or applicable  stan-
              dard, ng/J.

7.2.2  Relative Accuracy Audit Calculations - Example   data  from
an RAA on an S02/O2 CEMS are shown in Table 7.3.


TABLE 7.3  RELATIVE ACCURACY AUDIT DATA FOR S02 AND C>2  CEMS

Run
number
1
2
3
Avg
so2
ppm
500
505
510
	
so2
CEMS,,
ppm
475
480
480
	
°2
>
3.0
3.0
3.0
	
°2.
CEMSd,
3.1
3.1
3.0
	
so2
RM
ng/J
422.4
426.6
430.8
426.6
so2
CEMS ,,
ng/JQ
403.5
407.7
405.4
405.5
         RM, = reference method data, dry basis.

       CEMS, = monitor data, dry basis.

-------
                                            Section No. 3.0.7
                                            Date November 5, 1985
                                            Page 8

    The S02 and 02 OEMS data shown in Table 7.3 were corrected to
a dry basil using Equation 7-1.  The SO2 and 02 GEMS and RAA data
were converted to the units  of  the  applicable  standard  using
Equation 7-2.

    The accuracy (A) for the RAA is calculated using Equation 7-7.
          A = Cm " Ca x 100
                                      Equation 7-7
              405.5 - 426.6

                  426.6
              x 100 =  - 4.95%
where
          A = accuracy of the CEMS, %,
         Cm '
average CEMS response during audit in units
of applicable standard, and

average audit value of the three reference
method runs in units of the applicable standard.
7.2.3  Cylinder Gas Audit  Calculations - Example data from a CGA
on an SO2/O2 CEMS are shown in Table 7.4.


      TABLE 7.4  CYLINDER GAS AUDIT DATA FOR S02 AND C>2 CEMS

Audit
number
1



2




Reading
No.
1
2
3
Avg
1
2
3
Avg
S02
CGA
ppm
212
212
208
210.7
398
399
403
400.0
S02 A
CEMS,, Diff,
ppm %
218
219
225
220.7 4.75
409
416
414
413 3.25
°2
°2 A
CGAd, CEMSd, Diff,
5.0
5.0
5.1
5.03
9.1
9.1
8.9
9.03
5.2
5.3
5.2
5.23 3.98
8.9
8.9
8.9
8.90 -1.44

-------
                                          Section No. 3.0.7
                                          Date November 5, 1985
                                          Page 9

   CGA, = cylinder gas audit value, dry basis.

  CEMS, = average of the three monitor values, dry basis.

  The S00 and Q~ CEMS data shown in Table 7.4 were corrected to a
dry  basis  using  Equation 7-1.  The accuracy (A) for the GCA is
calculated using Equation 7-8.
              c  - C
          A =  m    a x 100                         Equation 7-8

                 Ca

            = 220.7 - 210.7 x 1QO  = 4>75%

                  210.7

where

           A = accuracy of the CEMS component, %,

          C  = CEMS component mean response for three values
               during audit with CGA in units of the appropriate
               concentration, and

          C  = audit value of the cylinder gas in units
               of appropriate concentration.

-------
                                            Section No. 3.0.7
                                            Date November 5, 1985
                                            Page 10
7.3  Reporting Requirements
    At the reporting interval specified in the  applicable  regu-
lation, a report of each CEMS accuracy audit must be submitted in
the form of a Data Accuracy Report (DAR).  One copy  of  the  DAR
must be included for  each  quarterly audit along with the report
of emissions required  under  the  applicable  regulation.   As a
minimum, the DAR must contain the following information:

    1.    Source owner or operator name and address.

    2.    Identification and location of monitors in the CEMS.

    3.    Manufacturer and model number of each  monitor  in  the
          CEMS.
                                                \

    4.    Assessment of CEMS data accuracy and date of assessment
          as determined by a RATA, RAA,  or CGA, including the RA
          for the RATA,  the  A for the RAA or CGA, the reference
          method  results,  certified  values  for  the  cylinder
          gases,  the  CEMS  responses,  and  the  CEMS  accuracy
          calculation  results.   If  the accuracy audit  results
          show the CEMS  to  be out-of-control, the CEMS operator
          shall report both the audit results showing the CEMS to
          be   out-of-control  and  the  results  of  the   audit
          following  corrective  action  showing  the  CEMS to be
          operating within specifications.

    5.    Results from the EPA performance audit samples.

    6.    Summary  of all corrective actions taken when the moni-
          tor was determined out-of-control.

    An example of a DAR form is shown in Figure 7.1.

-------
                                                      Section No. 3.0.7
                                                      Date September 23, 1985
                                                      Page 11
Period ending date 	  Year

Company name 	
Plant name 	  Source unit no.

CEMS manufacturer                          Model no.
CEMS serial no. 	,	 CEMS type (e.g., in situ)

CEMS sampling location (e.g.,  control device outlet) 	

CEMS span values as per the applicable regulation, S0_ ppm
  0_ 	 percent,  NO  	 ppm, CO- 	 percent
   ^     '                       X      "                £

I.  Accuracy  assessment  results (Complete A, B, or C below for each CEMS
    or for each pollutant  and  diluent  analyzer, as applicable.)  If the
    quarterly audit results show the CEMS to be out-of-control, report the
    results  of both the quarterly  audit  and  the  audit  following  the
    corrective action showing the CEMS to be operating properly.

    A.  Relative accuracy test audit (RATA) for 	
        (e.g., S02 in ng/J).

        1.  Date of Audit
        2.  Reference methods  (RM's) used 	 (e.g., Methods 3 and 6).

        3.  Average RM value 	 (e.g., ng/J, mg/dsm , or percent
            volume).

        4.  Average CEMS value 	.
        5.  Absolute value of the mean difference  IdI

        6.  Confidence coefficient  |CC|
        7.  Percent relative accuracy  (RA) 	 percent.

        8.  EPA performance audit results:

            a.  Audit lot number          (1)	  (2)	
            b.  Audit sample number       (1)	  (2)	
            c.  Results  (mg/dsm-5)         (1)              (2)
            d.  Actual value (mg/dsm3)*   (1) 	  (2) 	
            e.  Relative error*           (1)    	       (2)
*To be completed by the Agency.

           Figure 7.1  Example format for data assessment report  (DAR).

-------
    B.   Cylinder gas audit (CGA)  for

        1.   Date of audit
        2.   Cylinder ID number

        3.   Date of certification

        4.   Type of certification


        5.   Certified audit value

        6.   CEMS response value

        7.   Accuracy
                                     Audit
                                    point 1
                                               Section No. 3.0.7
                                               Date September 23, 1985
                                               Page 12

                                              (e.g.,  SCL in ppm).
                                         Audit
                                        point 2
                                                 (e.g., EPA Protocol 1
                                                 or CRM).

                                                 (e.g., ppm).

                                                 (e.g., ppm).

                                                 percent.
    C.   Relative accuracy audit (RAA)  for

        1.   Date of audit
        2.   Reference methods (RM's)  used

        3.   Average RM value 	
4.  Average CEMS value 	

5.  Accuracy           	

6.  EPA performance audit results:

    a.  Audit lot number         (1)
    b.  Audit sample number      (1)~
    c.  Results (mg/dsnr)*  _    (1)
    d.  Actual value (mg/dsnr)*  (1)
    e.  Relative error*          (1)
                                              (e.g.,  S02 in ng/J).
                                                 (e.g.. Methods 3 and 6)
                                          (e.g., ng/J).
                                                  percent.
                                                         (2)
                                                         (2)"
                                                         (2)"
                                                         (2)
                                                         (2)
*To be completed by the Agency.
                                  Figure f.l (continued)

-------
                                                       Section No.  3.0.7
                                                       Date September 23,  1985
                                                       Page 13
    D.   Corrective action for excessive inaccuracy.

        1.   Out-of-control periods.

            a.  Date(s)	
            b.  Number of days
        2.  Corrective action taken
        3.  Results of audit following corrective action.  (Use format of
            A, B, or C above, as applicable.
II.  Calibration drift assessment.

     A.  Out-of-control periods.

         1.  Date(s)
         2.  Number of days
     B.  Corrective action taken
                              Figure 7-1 (continued)

-------
                                          Section No. 3.0.7
                                          Date September 23,  1985
                                          Page 14
7.4  References
     I. Standards  of  Performance  for  New  Stationary  Sources:
        40 CFR  60,  Appendix  F  - Quality Assurance  Procedures,
        Procedure  1  - Quality Assurance Requirements for Gaseous
        Continuous Emission Monitoring Systems Used for Compliance
        Determination.

     2. A Procedure  for Establishing Traceability of Gas Mixtures
        to Certain National Bureau of Standards Standard Reference
        Materials.    Joint   publication   by   NBS    and   EPA,
        EPA-600/7-81-010.  Available from the U.  S. Environmental
        Protection  Agency,  Quality  Assurance  Division (MD-77),
        Research Triangle Park, North Carolina   27711.

     3. Traceability Protocol for Establishing True Concentrations
        of Gases Used for Calibration  and  Audits  of  Continuous
        Source Emission Monitors (Protocol Number 1).  June  1978,
        Section  3.0.4 of the Quality Assurance Handbook  for  Air
        Pollution  Measurement  Systems,  Volume  III,  Stationary
        Source   Specific  Methods.   EPA-600/4-77-027b.    August
        1977.  U. S. Environmental  Protection  Agency,  Office of
        Research  and  Development Publications, 26 West St. Clair
        Street, Cincinnati, Ohio   45268.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                               Page 1
9.0  CONTINUOUS EMISSION MONITORING (CEM)  SYSTEMS GOOD OPERATING
     PRACTICES

    Continuous emission  monitoring  (CEM)  systems are  required to be
installed in facilities specified by the EPA Standards of Performance
for New  Stationary Sources  (SPNSS)  and by  other Federal  and state
regulations.  The systems are used to continuously monitor the effec-
tiveness  of air  pollution control  techniques  and  to  determine if
source compliance standards are being met.

    This  section  of  Volume III  is  intended to  provide  guidance  for
technical personnel in air pollution control agencies and in industry
who are  responsible  for CEM programs.   Guidelines are  given  to  aid
agency  personnel  in  evaluating  operation  and  quality  assurance
practices associated  with permanently  installed  CEM systems.    The
guidelines  may  also  be  useful  to  operators   of  CEM systems  in
developing quality assurance and quality control procedures that meet
agency minimum requirements.   Section 3.0.9  does not address the  use
of  continuous  monitors  in  mobile  testing  vans  or  as  portable
compliance monitors.   However, much of the information presented here
is relevant to these applications.

    CEM  systems have  been  developed to  monitor pollutant gases, such
as SO2 and  NO,  and the  so-called diluent  gases, C02 and 02, present
in the exhaust gas streams of  combustion  sources.  Systems have also
been developed to  monitor  flue-gas opacity.   A  system is defined as
the  total  equipment  required  for  the   determination   of  flue-gas
opacity,  a  gas concentration,  or  the  emission  rate.    A  system is
normally  composed  of a  sample interface,  the  pollutant and diluent
analyzers,  and the data recording subsystem.  The  system is used to
generate emission data that are  representative of the  total emissions
from the facility.

    The  sample interface is the  portion of the monitoring system that
protects  the  analyzer  from  the  effects  of  the  effluent.    In
extractive  systems,  the  interface  consists  of   the  probe  assembly,
sampling lines, and  conditioning subsystems.   The sample is normally
taken from  a  single  point in  the stack or  duct and then transported
to  the   analyzer.   A  conditioning system  is  often used  to  remove
particulate matter from the sample and  to dry  the  sample  before it
enters the analyzer.

    In-situ  monitors have been  developed  to  measure the  stack  gas
concentrations, without  transporting the  gas  itself.   Gas measure-
ments are  made either  at  a point  or along  a path  of  known length
within  the  flue.   For  in-situ  path monitoring, the  interface  may
consist  of  optical windows  and  blower assemblies  used  to  keep  the

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                               Page 2
windows clean.  For point  in-situ  designs,  it  may consist of ceramic
thimbles  and support  housings.    These  different  approaches  taken
toward the measurement  of  effluent gases will be  discussed later in
this section.

    The  SPNSS  require  data  obtained  from  a  CEM  system  to  be
representative,   accurate,  and  precise.     In  contrast  to  EPA
certification procedures  for ambient  air monitors,  source emission
monitors  are not  categorically  approved by  model  or  manufacturer.
Instead,  installed systems  are  approved  on  a   case-by-case  basis
through the  procedures  established  in  the Performance Specifications
for  Continuous  Emission  Monitoring  Systems  in   Stationary  Sources
(40CFR60 Appendix B).i  After an installed monitoring system is found
to meet these specifications, it is expected that it will be properly
maintained at the same or better level of performance.

    The  proper  operation  and  maintenance   of   a   CEM  system  is
imperative if the  data  are to be used for  regulatory purposes.   The
responsibility for  the  system lies with  the owner,  and  in general,
ownership lies with the plant or industrial  facility.  The generation
of  valid  data   from   a CEM system   through  proper  operation  and
maintenance  procedures  must  therefore come  from plant  personnel or
through services contracted by the plant.   It  is,  however,  the plant
personnel or their contractors who will actually operate and maintain
these systems.

9.1  CEM Operation/Maintenance Programs  - Levels  of  Quality Control

    A  maintenance program  for   a  CEM  system  should be  part  of  a
larger,  plant  instrumentation  quality  assurance   (QA)   program.
Quality control practices within the QA  program  are those activities
performed to assure that accurate and precise data are generated from
the monitoring  system.   Daily  operation checks,  preventive  mainte-
nance routines,  and audits are quality control  activities that can be
used for this purpose.

    There are four levels of quality > control  that  should  be estab-
lished for a CEM system:

    Level 1. Operation  Checks   (daily  checks,  observations,   and
             adjustments)
    Level 2. Routine Maintenance (periodic preventive maintenance)
    Level 3. Performance Audits
    Level 4. Corrective Maintenance

    Operation checks  are  performed on  a  routine  basis,  generally
daily,  to see  that  the  equipment  is  operating  properly.    These

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                               Page 3
procedures will include daily zero  and  calibration checks, checks of
reference  signals  from  control  panels,  and checks  of  flow rates,
pressures, vacuum levels.

    Routine   maintenance   is   performed  at   regular   intervals.
Activities include the replacement of filters, lamps, motor bearings,
or other  parts.   Detailed  service  checks of electronic  and optical
systems may also  take  place at  this time  to uncover incipient prob-
lems in  the  instrumentation.   Depending on  the  system,  the  replace-
ment and check intervals may vary from 30 days to a year or more.

    Performance  audits,  which  provide  a  check  of  the  system's
operation, identify problems,  identify the need to improve preventive
maintenance  procedures,  or alert   the   operator  to  the  need  for
corrective maintenance.

    Corrective  maintenance  is   performed  to bring  the  monitoring
system into operation after a breakdown  in the  system occurs.  It is
also termed nonroutine maintenance,  the  unscheduled  need  to repair a
faulty system.

9.2  Gas GEM Systems - Operation Practices

    The  day-to-day  operation  of a  GEM  system is  not  difficult once
the instruments are  turned on  and  operating properly;  generally all
that needs to be done  is  to periodically check  the zero and the span
of the instruments in the system.  This check may be conducted either
manually  or   automatically by  using  calibration gases   or  optical
filters.  However, routine and corrective maintenance practices vary,
depending upon the methods  of analysis  and the  overall  design of the
system.   For  this  reason,  it  is important  to understand  the special
demands  of different  monitoring  systems.   Extractive systems  have
different maintenance  requirements  than  in-situ  systems.   Within the
categories of  extractive  systems,  or in-situ systems,  the different
types of  analyzers will  require  servicing dependent  on the principle
by which they analyze the pollutant.

    To  help  understand maintenance  requirements, this section  will
present  an overview  of the various  analysis principles  used in the
commercial systems.2  Table 9.1  summarizes these principles.

9.2.1    Extractive  Monitoring  Systems  -  This   section   will  first
discuss  the  design of extractive systems  and then  the  various  ana-
lytical techniques used to measure the gas concentration.

9.2.1.1   Extractive  System Design.   A complete extractive system
consists  of  a sample  probe and  conditioning system,  analyzer,  and

-------
                                                       Section No. 3.0.9
                                                             Date 6/1/86
                                                                  Page 4
              TABLE 9.1.  PRINCIPLES OP  DETECTION USED IN
                      CONTINUOUS EMISSION MONITORS
     Extractive System
               In-Situ  Systems
     Gaseous  Emission
         Monitors
      Gaseous Emission
          Monitors
   Opacity
   Monitors
Absorption Spectroscopy
   Nondispersive Infrared
   Differential Absorption
Luminescence Methods
   Fluorescence
   Chemiluminescence
   Flame Photometry

Electroanalytical Methods
   Polarography
   Electrocatalysis

Paramagnetic Methods
Absorption Spectroscopy
   Nondispersive Infrared
   (Gas filter-correlation)
   Differential Absorption
   Second Derivative
     Spectroscopy

Electroanalytical Methods
   Electrocatalysis
    Visible
Light Scattering
  and Absorption
data recording  system.   Sampling probes and  conditioning  systems are
today  commonly   purchased  from  the  analyzer  vendor  rather  than
assembled  from miscellaneous  parts by  plant  technicians.   Also,  a
number of  companies specialize in marketing  hybrid systems  (complete
extractive  systems  composed  of  components   supplied  by  different
vendors).

    There  are two approaches  taken in extractive system design.   One
is  to  condition  the  gas  near  the  analyzer;  the  other, to  condition
the gas  as close as  possible to  the stack  or  duct.   In  the first
approach,  a probe is inserted  into the flue  gas  and the gas is drawn
through  a  coarse  particulate filter into a heated sampling line.   The
sample   line   may  extend  to  over   60 m  to   a   control   room  or
environmental  enclosure,  where the  gas  is  conditioned.   The  condi-
tioning  system  cools the  gas  and removes  water vapor by some type of
refrigeration, dilution, or permeation device.   Usually  a  fine filter
is  placed  just before  the  analyzer to  prevent  small particles  from
entering the  analyzer.    Diaphragm pumps,   rotary vane  pumps,  or air
aspirators  are used  to  transport  the sample  from the  probe  to the
analyzer.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                               Page  5
    In the  second  approach of sample extraction,  gas is conditioned
at the  stack  or duct.   Filters,  chillers,  or dilution  systems are
located  at  the  sampling  site,  and  in  the  case  of  some  dilution
designs,   the  probe  itself  does  the  conditioning.    This  approach
allows a low moisture sample to be transported to the analyzer.  Long
sections of heat-traced or insulated lines may therefore be avoided.

    Calibration  gases  are  used   in  both  approaches  to  check  the
performance of  the  system.  The gases  are injected  as  close  to the
probe as is technically  feasible.   Also,  blow-back devices are often
installed to  clean the  coarse  particulate filters.   As  the  system
operates, these  filters may  eventually plug  up.    A burst  of  high
pressure air  "blown back"  through the  filter reduces  plugging and
provides for continued operation.

    Extractive systems are normally  constructed  from components that
are  familiar  to  plant  mechanics.   Valves,   filters,   tubing,  tube
fittings, solenoids,  etc., are  commonly  encountered.   These  compo-
nents must be maintained if the system is to provide continuous data.

9.2.1.2  Extractive Analyzers  -  Spectroscopic Absorption Techniques.
Two  basic  absorption  spectroscopic   techniques   are   utilized   in
commercially  available  extractive  analyzers:   (1)   non-dispersive
infrared spectroscopy and  (2) differential absorption spectroscopy.

    Nondispersive infrared  spectroscopy utilizes  infrared  light  in a
limited  range  of  the  electromagnetic  spectrum.    The  light  is not
scanned or "dispersed" as with scanning laboratory spectrometers.   In
general,  the light is  filtered to  select  light wavelengths that will
be absorbed  by the molecules  that are  to be  measured.    The light
passes through a gas  cell that contains  the  flue  gas extracted from
the stack.   A portion  of  the light  from the lamp passes  through a
cell  containing  a  reference  gas  that  does  not  absorb  the  filtered
light.   A  detector  senses  the  amount  of light  absorption  in the
sample cell relative to  the signal from the reference cell.   Through
proper calibration,  the detector  responses are  electronically  con-
verted to pollutant concentration  readings.   A variant  of this tech-
nique, called  gas  filter  correlation spectroscopy,  uses  a reference
cell  that  absorbs  100% of the  light  in  the molecular  absorption
region of the pollutant.

    Infrared analyzers  have been developed to measure  gases  such  as
S02/  NO,  NO2f  HC1,  C02, and  CO.   The  commercially available monitors
differ primarily  in  the  design  of the  detector  and  the  level   of
rejection of interfering gases.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                               Page 6
    Differential absorption spectroscopy  also  takes advantage of the
characteristics of molecules  to  absorb  light of certain wavelengths.
Instead of  using  a sample cell  and  reference  cell as  do  the common
infrared systems,  differential absorption spectrometers use a measur-
ing "wavelength" corresponding to a  region  of  the spectrum where the
molecule absorbs light  energy.   The  reference  wavelength corresponds
to a region where there is little or no absorption.

    Most  extractive  differential absorption systems  operate  in the
ultraviolet  (UV) region of  the spectrum, although  it  is  possible to
use the technique  in the  infrared region.   The gases  may be measured
hot in the  UV  without  removing water vapor, although  it is generally
advisable to dry the sample.

    S02 is  commonly  measured  using  this  technique.   A technique has
also been devised to measure flue gas NO by  injecting oxygen into the
sample chamber, sealing it,  and monitoring  the production of NO2 from
NO at an N02 absorption wavelength.

9.2.1.3     Extractive   Analyzers   -   Luminescence   Techniques.
Luminescence is the  emission  of light  from a  molecule or  atom that
has been  excited  in  some manner.  Three  luminescence techniques are
used in the  field  of source monitoring: (1) fluorescence,   (2) chemi-
luminescence, and  (3) flame photometry.

    Ultraviolet fluorescence  is used  to  measure  SO2-   Ultraviolet
light in the region of  210 nm is used to excite an  S02 molecule.  The
molecular excited  state persists for a  few nanoseconds, during which
time some  of  the  energy is  lost  in  vibrational  transitions.   The
molecule eventually  returns to its   unexcited  state with  the release
of light  at a  longer wavelength (near  350  urn).   This  light is then
detected by a photomultiplier  tube,  resulting in a measurement of the
SO2 concentration  in the sample gas.

    Fluorescence monitors can  be affected by changes  in the flue-gas
composition  (%O2,  %C02).    This  is  caused  by the de-excitation  of
excited SO2  molecules   through the  process  of  quenching.    For this
reason, fluorescence analyzers are most successful  in flue-gas analy-
sis when they are coupled with a dilution system, thereby providing a
relatively constant background composition.

    Chemiluminescence is  used  in flue-gas analysis to measure NO and
NO2  concentrations.     In  this   application  of  chemiluminescence,
excited NO2 molecules  are produced  by  reacting ozone  with the flue
gas NO.    The  excited  NO2  product   (NO2*)  de-excites  to  its ground
state with  the release  of light energy.  The light  is measured with a
photomultiplier tube.   Quenching effects also  occur  in this method,

-------
                                                    Section No.  3.0.9
                                                          Date 6/1/86
                                                               Page 7
but dilution of  the  sample  through the introduction of  the reactant
ozone gas stream minimizes the effect.

    Since the  light  is produced only  through the reaction  of  ozone
with NO, NO2 must  first be  reduced to NO before  it  can  be measured.
A catalytic  reaction  chamber  is used  when  a measurement  of  both NO
and N02 (NOx) must  be  obtained.

    Flame photometry  can  be  used  to  measure compounds  that contain
sulfur.  In this technique, the compounds are "burned"  in a hydrogen
flame, leading  to the formation of  excited diatomic sulfur molecules,
82*.   The conversion  of the  high  energy  82* molecules  to the  lower
energy  ground  state,   82,  occurs  with  the  emission of  light.    The
intensity of this  light is  measured and related  to the concentration
of  sulfur species  in  the  sample.   The  flame  photometric method does
not  discriminate  between  different  sulfur-containing  compounds,  so
scrubbers or gas chromatographic columns may be required if more than
one species  is  present in  the  sample.

9.2.1.4   Extractive  Analyzers - Electroanalytical Techniques.   Two
principal electroanalytical techniques have  been developed  for  the
measurement   of  flue  gases.   These  are polarography   and  electro-
catalysis.    A  clean,  dry  sample  must  be   supplied  to  an  analyzer
ope.rating  by   the polarographic   method.     The   electrocatalytic
technique can,  however, be applied  to  both extractive  and in-situ
measurement  methods.

    Polaroqraphic  analyzers  are,   basically,   diffusion-controlled
electrochemical  cells.     The  cells  are   constructed  much   like
batteries,   with  a  sensing  electrode,  electrolyte,   and  counter-
electrode.    The  main  difference  is   the  addition of   a thin-film
membrane, through  which the  pollutant  must  diffuse to  initiate  the
electrochemical reactions and current  flow.   The  current across  the
cell  is  proportional  to the rate  of diffusion of  the  pollutant into
the cell and is also proportional  to the pollutant concentration.

    Polarographic analyzers have been developed to measure gases such
as  S02,  NO,  02,  and  C02.   Different choices  of  electrodes and  elec-
trolytes are made  for  each  gas.   As with batteries,  the  electrolyte
will eventually be consumed, and the cell will need to  be  replaced or
recharged.

    Electrocatalytic   analyzers  have  been   developed   for  the
measurement  of  ©2  and  S02.   This  technique  uses  a solid  electrolyte
instead  of   liquid  electrolytes   generally  associated  with
electrochemical cells.  A platinum  film, coated on the  solid surface,

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                               Page 8
catalyzes a  reaction that  allows molecules  to migrate  through the
solid and generate a measurable flow of electrons.

    In  oxygen  electrocatalytic  analyzers,  a  zirconium  oxide disc;
coated with a thin film of platinum, is heated to 850°C.  A reference
gas of about 21%  oxygen  is  maintained on one  side  of  the solid, and
the sample gas  is on the other  side.   Oxygen  ions  are generated at
the  platinum surface  and  then  migrate  through   vacancies   in  the
heated, solid electrolyte.  Electrons  are  released  in  the process as
the system attempts to equalize the oxygen concentration.

    An  electrocatalytic  analyzer  has  also  been  developed  for  the
measurement of  SO2.   This system uses a potassium sulfate crystal and
requires   the   simultaneous   measurement  of  the  sample   oxygen
concentration.

9.2.1.5   Extractive  Analyzers  - Paramagnetic  Techniques.   Oxygen
exhibits  paramagnetic  behavior   by  being  attracted  to a  magnetic
field.   This behavior  has  been  utilized in  the  design  of   several
different types  of  extractive flue-gas  analyzers.   For  example, in
thermomaqnetic  oxygen analyzers,  a magnet causes 02 to flow through a
tube and cool a resistor.  The resistance is then related to 02   con-
centration.  In maqnetodynamic systems,  02  disturbs a  magnetic field
around a torsion  pendulum,  and  in paramagnetic pressure analyzers, a
magnetic field  causes a pressure imbalance that can be measured.

9.2.2  Recommended Maintenance -  Extractive  Monitoring  Systems

9.2.2.1   Operation Checks  (Daily Checks).    Operation  checks  of an
extractive  monitoring  system  should  be  performed  each  day by  a
qualified and  trained  instrument operator.   The operator  should be
familiar with  the system and be  able to  recognize a  problem  from
discrepancies  found  during the  operation  check   procedure.     Many
extractive monitoring  systems  are designed to automatically   perform
daily  zero  and  calibration  checks  and internal self  checks   without
operator intervention.   Unfortunately,  this can reduce  the level of
operator attention to  the system.  Small problems,  as  a result, may
go undetected and very quickly lead  to  large  problems.   On the other
hand,  "intelligent" systems, which  monitor  key system  parameters and
report  out-of-control  conditions at  remote  panels,  have  helped to
alleviate such  situations.

    A  daily operation check  of an extractive monitoring  system should
start  with a  check of  the  strip  chart  record  and/or  other   data
recording devices.   The  operator should  mark the  exact  time on the
chart  for calibration  purposes,  and write down the  date,  his or her
name,  and  the  chart recorder settings.   This should  all be   written

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                               Page 9
directly on  the  chart.   The  paper  in  the recorder  and  printer  (if
applicable)  should be checked to  see if the supply is sufficient  for
the next 24-h run.

    Indicator lights on the system or monitor control panel should be
checked next.   It is advised that  a record of  the  system status be
recorded in ink in a hardbound logbook.  All maintenance,  unscheduled
repairs, or system modifications  should be described in the  logbook.
This book will serve as an  invaluable tool in tracking the long-term
performance  of  the  system  and  will  enable  other   technicians  or
servicemen to become familiar  with the system.

    The system  indicator  lights  will notify  the operator of out-of-
range conditions occurring in  the system or of other problems felt to
be important by the system designer.   If a problem occurs, it  should
be attended  to immediately since  subsequent  data will  be otherwise
suspect.  Some  systems contain reset  buttons,  installed   to  override
the  indicator  lights.    These  should  not  be  used  until  after  the
problem is resolved.  The values of other systems indicators, such as
vacuum or pressure  gauges,  sample flow  rates,  and  lamp and  detector
reference levels  (if  applicable),  should  also  be recorded  at this
time.

    A  calibration check  should  be  performed  next.3   This  involves
injecting a zero  (or low-level)  gas and a high-level gas  (calibration
gas)  into the  sample  line.    It  is  recommended  that  the  gas  be
injected  at  a  point   where  as  many   of  the   conditioning  system
components as possible can be checked.   In some systems,   this  can be
done  at  the probe  itself,  providing  the  advantage  of  checking  the
system for sample line losses.

    Gas  injected  from  gas  cylinders   may pressurize   the  system.
Consequently,  the flow rate  of  the  cylinder  gas into  the  analyzer
sample cell may differ from the flow  rate of the  extracted stack gas.
Also, cylinder gas  is  dry gas;  it does  not contain  moisture.   These
two  factors  may  adversely affect  the  calibration  process  if flow
rates and moisture content of  gases entering the  analyzer  sample cell
are not similar.  It is also  important  to note that if the system is
pressurized, leaks  in  the system may not be detected.   Ambient  air,
will not enter  the  system as  it otherwise  might  if  a vacuum is used
to draw a sample into the analyzer when leaks are present.

    Gases used for the daily checks  should  first  be validated against
certified calibration  gases or  be  certified  themselves.   Certified
gases  should  have  their  concentration   established  through   EPA

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 10
Traceability Protocol No. 1.3  Using gases of uncertain concentration
can result in gross miscalibration of the system.

    The monitor should first be  checked  with the zero (or low-level)
gas, and the instrument  reading  should be  noted  in the logbook.  The
high-level  gas  should  next  be   injected  into  the  system  and  the
reading likewise noted.  The readings  should also be recorded  appro-
priately on  the  strip chart record.   The differences  noted between
the cylinder gas  value and the  monitor  readings are  used to  assess
the low-level and high-level calibration drift.

    The   instrument  operator  may   not   have   to  "rezero"  and
"recalibrate" the  system every  24  h  when  the  values  are  checked.
Small values  for  drift  may  be  due  merely to system  noise.   It is
recommended as a minimum that  the system  be  adjusted  when the drift
exceeds  twice  the  limits  of  the drift  performance  specification.4
For  example,  if  the  performance  specification, is  2.5%  for  an
instrument span value  of 1000 ppm,  adjustments  should  be made when
drift exceeds 50 ppm.  For systems  with  lower span values, the drift
tolerance will be accordingly  less.   (The  span value is  given  in the
Code  of  Federal  Regulations   for  source  categories  affected  by
continuous monitoring regulations.  The span value is defined as "The
upper  limit  of  a  gas  concentration  measurement  range   that  is
specified for affected source categories in the applicable subpart of
the regulation." i)

    The operator should  record values  from the  instrument meter, the
strip chart, and digital printer.   If  a  microprocessor controller is
used  to  check   and/or  adjust  monitor  data automatically  to  the
appropriate values,  it must be  programmed to record  the unadjusted
values first.  If a  strip chart  recorder  is  used in conjunction with
the  microprocessor,  the  system  should  be  programmed  so  that  the
adjusted values will appear  on both the strip chart and the printer
output.  This may be difficult since the microprocessor adjustment is
often only done numerically by the program, i.e., the analyzer  itself
is not physically adjusted.   In  such a case, the meter  readings and
strip chart readings may differ  significantly from the microprocessor
output.  Data interpretation in such cases may become difficult.

    To assist in performing  the  daily  operation  checks,  a data sheet
has been provided in Figure  9.1.  The figure is  meant to serve as  a
guide  for  the  inspector  or  operator  in  developing a  data  sheet
applicable to a specific system.

9.2.2.2   Routine Maintenance  (30-day  Checks).    Routine maintenance
should be initially  performed  on an extractive system at least every
30 days.  With  experience/  this  time period  can be either  increased

-------
                                                  Section  No.  3.0.9
                                                        Date  6/1/86
                                                           Page  11
Example Format for Extractive Gas Monitoring System
Daily QC Check Sheet
Plant f flJvCl/Tder /^ m^ fi^UW Date S/.ao/5?6 Time f .' 3-
Analvzer I.D. MgQ<;i;itf rrh +*4-7tj
Span Value IOQO ooivi
Calibration Gas Value c]Li'S
Zero Gas Value (air, N2, oth
Hours Ooeratinq in Period
er) a\C

Part 1 Indicators
Indicator Lights/Gauges
Sample pressure/vacuum
Sample flow
Lamp
Detector
Name Joe. JonCS
Phone 919 - 5HT- 0*6,3
Zero Offset value ^>O ^>prvi
Date Certified q/i<-f/S3

n


Paper Status: Strip Char
Printer 34

Status
IO psi.
0.3SJL/^
OK
OK
Part 2 Calibration Check
Unadjusted Readings
Zero (low-level) gas
Calibration (high-level) gas
Stack Concentration
Time
q ; 50 a,/^
10: co
10'. 10
Part 3 Zero and Span Adjustment (if
Control limit i 50

Adjusted Readings
Zero (low-level) gas
Calibration (high-level) gas
Stack Concentration
t (5K
hrs -rp(i'«««te»

Problem/Action taken
Louj/ (Boost "fct> 1 2- ps i_
^
-
-

Meter
!(.
Mff
^^8
Strip Chart
If*
BUS
5HS
Digital Printer
at

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 12
or decreased, and  depending  on the system,  maintenance intervals of
varying frequency may be established for subsystems or  for individual
components.

    In extractive monitoring systems,  most  maintenance lies with the
sample  conditioning   systems   rather   than  with  the   analyzers.
Particulate  matter and  water vapor are usually  removed from the gas
stream before the gas  reaches  the  analyzer.   The filters that remove
the particulate  matter  must  be  periodically  cleaned or   replaced.
Condensed water  in  condensing-type moisture  removal systems must be
drained.

    The  plumbing  associated  with  extractive  systems  is   prone  to
corrosion and  leaks;   therefore,  the  system should be periodically
checked for  leaks.  Fittings,  valves,  and  gas regulators should also
be checked.   Solenoid valves  have  been  commonly  used  to  automate
extractive systems.  These valves  are  prone  to failure and should be
checked frequently to  ensure they move  freely and on  command.   The
use of motorized or  air-activated rotary valves  instead  of solenoid
valves may also help to reduce  the frequency of valve failures.  Care
should be taken  to avoid over-design or  over-automation of a system.
The more valves there are,  the  more valves there are to check.  Spare
valves should be kept in the parts inventory.

    Electrical cables  and  heat-traced  lines should also  be checked
frequently.     In  a   plant  environment,   damage  can  occur  from
construction  projects   or   through normal   plant  operations.    The
ambient atmosphere, particularly near  flue-gas leakage or stack down-
wash areas,  may  cause  electrical  insulation to  deteriorate rapidly.
Acid gases  circulating near the  stack  may  corrode both electrical
fittings and the plumbing of the extractive system.

    The pumps and chillers used  in extractive systems  work 24 h/day.
At some time, the motor brushes will wear  out, a pump diaphragm will
break, or a part  will require  oiling.    Rather than  treating  such
events as  problems, they  should  be   anticipated  by  establishing  a
regular schedule of overhaul and maintenance.

    The  overall  cleanliness   of  an  extractive  system  is  also
important.  The  particulate  matter in the  flue-gas  can migrate into
unexpected places.  If a  system is located  outside,  near the stack,
sensitive components should be installed  in dust-free  cabinets.   The
system should be cleaned if fly  ash  settles  on it,  and  in no case
should cabinets  be opened when fly ash is  circulating  in the ambient
air.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 13
    The  extractive  analyzers contain  components which  have  limited
lifetimes.   Lamps and bulbs generally have given performance periods.
They should  be  replaced before this  period  is up,  since a weakened
bulb  can  often  produce  spurious   results.    Analyzers are often
designed to operate best over a given  range  of lamp intensity.  When
the intensity drops too low, the detector will not be able to  respond
as accurately  as it  should to the  incoming signal.   Detectors may
also have to be replaced, but this is not common.

    Many instruments have test points on the back panel or on  circuit
boards.    These   test  points  are checked  with  a  voltmeter  or  an
oscilloscope to  indicate  certain limiting  values.    The instrument
operator should perform these electronic tests routinely  to check for
electronic integrity.

    Figure 9.2 gives an example of a checklist designed for a  routine
maintenance  procedure.   Again,  this list is  suggested  to  help the
inspector or operator design his or her own list.  It is  not uncommon
for  instrument  operators  to spend  a year  or  more in  designing   a
system maintenance  schedule.   Vendor  instruction manuals  are often
lacking  in this  regard,  so  points  of  maintenance may  have to be
determined through  experience.   The system  logbook  is an invaluable
tool in developing such a schedule.

9.2.2.3  Performance  Audits.   Performance audits should  be conducted
on  extractive   monitoring   systems  at  appropriate  intervals.   EPA
40CFR60  Appendix F - Quality Assurance Procedure I5 requires  that an
audit be performed at least once every quarter for monitoring  systems
used  for  determining  compliance with   emission  standards.    This
frequency  is  recommended  to  identify  CEM  systems  that   may  be
generating biased results.

    The  performance audit  is essentially  an  independent  check  of the
system,  and  can vary, depending  upon  the resources of  the  owner of
the system.  For CEM systems installed to demonstrate compliance with
emissions  standards,   EPA  requires   an   audit  at   least once each
quarter, using one of the followingS:

    •  Relative  Accuracy Test Audit  (RATA)

       A repeat   of  the  relative  accuracy   test  procedures   as
       defined  in Appendix  B Performance  Specifications.1

    •  Cylinder  Gas  Audit (CGA)

       Challenging  the  monitoring system with  cylinder gas   of
       known  concentration  (certified gases).

-------
                                                          Section No.  3.0.9
                                                                  Date  6/1/86
                                                                      Page 14
           Example Format for Extractive Gas Monitoring System
                     30-Day  Maintenance Check Sheet
Plant
Unit
       ~B&\\f!r 1*3. - OutU»±-
 System I.D.
 Analyzer I.D.
 Analyzer I.D.
 Analyzer I.D.
                          4*3^
Date_
Name
Phone
                                           tf3.7 /?5
                                                       Time
                                              - 543 - 8k 3
                            Gas_
                            Gas.
                            Gas
                         Required Haintenance Checks
Extractive System
Probe filter
Fine filter
Condensation system drain
Heat trace continuity
Pump - bearing noise
Plumbing leak check - vacuum
- pressure
Cable integrity
Cleanliness
Corrosion levels - probe
Solenoid performance
Regulator pressures
Air-operated valves
Air purge/blouback l
Zero gas
Calibration gas
Status
o£^^xSWete*"fc
&ra.u
JDrruV^ p^pert^
L\'*\e. (juarm -to-toycK
S(-v=,k-t ITLSpl "TO
NA
"Pressurised, 3o psi.
3^t£taOUSe. te^W-t. CQUJ&P&
pcSHcl-e aett'ii^ o»-v ccUsiei
Ply OA'n OLCCumuloM^ f><\.
Wo rust
Set£S^i IS^

(pCOp^x.
1 1 CO (XSL.
S&C p9L
(pao p»L
Action
)VlOK>e.
'R*^SJ"U*c^tf "eXt
^TXn^odL i^iCL'^OCLLU.A
^LOCXit^iA^ C^QX'L, i
Mo^e
C^ec^t- ^-o^t. 36 £^.(Xys


OleD«_*^ ^d
C^ecL*\fid. — c*Tdc»"tcX.
p^OttfC^t ^€. GO^G^
—
T^fA^t-d

Mo*^€
Ncv^C,
Wone
NJfcnt
          Analyzers
Lamp
Sensor
Test points
Chopper motor
Optical window status
U)o noise
C\«o^
-

j\ct±. to l.foO o-i^
-
-
                                       Jo
 Operator Signature
                          Date
                                    Supervisor Signature
                                                              D
                                                               ate
Figure  9.2.   Example format for extractive gas monitoring
            system 30-day  maintenance  check sheet.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 15
    •  Relative Accuracy Audit  (RAA)

       An  audit  similar to  the RATA except that  only three sets
       of  measurement  data  (instead  of  nine) are taken.

    Appendix F requires that at  least  one of  the quarterly audits be
a RATA, and  either the CGA or RAA can  be used  for  the other  three
quarters.    If  the relative  accuracy  between  the  audit  and analyzer
values exceeds 20% for the  RATA,  the  instrument  is  viewed as  being
out  of control.    If  the   relative  accuracy  between the  audit and
analyzer  values  exceeds  15% for  the  CGA or  RAA,  the  instrument is
viewed  as being  out  of   control.    A  RATA,   CGA,  or  RAA  must be
conducted after repairs are made to out-of-control systems.

    Additional techniques  can  also be  used  during  an  audit.   For
example, portable gas monitoring instruments can be used to check the
stack  gas concentrations  rapidly.     Although  the  portable   monitor
itself  may  not  meet  the   performance specifications  that  the CEM
system meets, it  can give valuable information during an audit.

    The  main idea behind  the performance audit  is to  provide an
independent  assessment of  the  monitoring system accuracy.     Daily
calibration  drift  determinations  and  routine  maintenance   do not
necessarily  guarantee that  data  will be accurate.   An  independent
assessment  using  an   appropriate  auditing technique can,  however,
provide  an  indication of   data  validity.    Figure  9.3   gives the
Appendix F example format for an audit "Data Assessment Report."

9.2.2.4    Corrective  Maintenance  (Problems   and  Troubleshooting).
Maintenance problems with extractive monitoring systems  usually  occur
in  the  gas  transport and  gas  conditioning  components.    Valves,
fittings,  tubing,  and  filters  in   the  presence  of  acid  gases,
submicron particulate  matter,  and continuous  vibration are likely to
have  limited life  unless   they  are routinely  maintained.   Lack of
routine maintenance or lack of foresight will result in the need for
corrective maintenance.   The need  for corrective maintenance can be
avoided by  establishing  good  quality  assurance  and quality  control
programs.

    The  extractive  system  gas  analyzers normally  will have few
problems  unless  they  are located  in  a severe  environment or if the
gas  conditioning  system fails.   If   the  conditioning  system fails,
acid gases can condense in  the sample  cell and  particulate matter can
settle  in the  system to  plug  the probe  or  sample lines,  or the
analyzer itself.

-------
                                                                      Section No.   3.0.9
                                                                               Date  6/1/86
                                                                                     Page  16
                 Example Format for  Data Assessment Reports
Period Ending Date   3 3o
                                                                  Year  I9SS
Company  Name  Acme. fkwie.f
Plant Hame
                                                    Source Unit Ho.
                                                    Model Mo.
CEM System Manufacturer  M'e£LSUf&-hsr>\
CEM System Serial Ho.   A 8106.3	     CEM System Type (e.g., in-situ) £x.4racH-w g.
CEM System Sampling Location  (e.g., control device outlet)  &SP  Out-let-
CEM System Span Values, as  per  the applicable  regulation, SO2
     O2  	 percent, HOx 	  ppm, C02 	
                                                                   loon
                                                                          ppm,
                                                                  percent
  I.  Accuracy Assessment Results.  Complete A,  B, or C below for each CEN  system or
      for  each pollutant and  diluenc analyzer,  as applicable.) If the quarterly audit
      results show the CEM System to be out of  control, report the results  of  both the
      quarterly audit and the audit following the corrective action showing the CEM
      System to be operating  properly.
      A.  Relative accuracy  test audit (RATA)  for
         1.   Date of Audit
         2
         3
         4
         5
                                                    (e.g., S02 in ng/J)
            Reference methods  (RMs) used 3onA(f (e.g.. Methods 3 and 6)
            Average RM value -434-.(a  (e.g., ng/J, mg/dsm, or  percent volume).
            Average CEM value  H5 I.a ng/J .
            Absolute value  of  mean difference  Idl l(c,.T3	•
            Confidence coefficient ICCI  3.~l• 3D	.
        7.  Percent relative accuracy  (RA)   IO. I 3
        8.  EPA performance audit  results:
            a.   Audit lot number            (1)
            b.   Audit sample number         (1)
            c.   Results (mg/dsm3)           (1)
            d.   Actual value (mg/dsm3)*     (1)
            e.   Relative error*             (1)
     B. Cylinder gas audit (CGA)  for
        1.  Date of Audit
                                                            percent.
                                                -C.1*
                                                              (2)
                                                              (2)
                                                              (2)
                                                              (2)
                                                              (2)
                                                                        . 3
                                                        Je.gr.,  SC<2  in  ppm)
                                 Audit point 1
        2.  Cylinder ID number
        3.  Date  of certification
        4.  Type  of certification

        5.  Certified audit value
        6.  CEMS  response value
        7.  Accuracy

*  To  be completed by the Agency
                                                Audit point 2
                                                         L. 3.01
                                      t 5
                                                            (e.g., EPA protocol 1 or
                                                            CRM).
                                                            (e.g., ppm).
                                                            (e.g., ppm).
                                                            percent.

                                                                         (continued)
Figure  9.3.   Example  format  for  data assessment  report.

-------
                                                                   Section  No.  3.0.9
                                                                            Date  6/1/86
                                                                                  Page  17
     C. Relative accuracy audit  (RAA) for  SOja.  in  noi jJ~ _ (e.g.,  SO2  in ng/J)
        1.   Date of audit
        2.   Reference methods (RMs) used 3aifdla (e.g. . Methods 3 and  6)
        3.   Average RM value 32.%. 4  (e.g., ng/J).
        4.   Average CEM value  A^3 . 8 _ .
        5.   Accuracy _ 3(5 % _ percent.
        6.   EPA  performance audit results:
            a.   Audit lot number           (1)      ^fe/6"        (2)
            b.   Audit sample number         (1)      'lOXZ         (2)
            c.   Results (mg/dsn>3)           (1)      jt. fc        (2)
            d.   Actual value (mg/dsm3)«     (1)      12£. &        (2)
            e.   Relative error*            (1)    ±(*.°( 4        (21
     D. Corrective action for excessive inaccuracy.
        1.  Out-of-control periods.
            a.   Date(s)    1/»iJtS- 7/3 1 /Sf
            b.   Number of days _ |? _ .
        2.  Corrective action taken  1$-ep\aLB\  [CLrtf>  77 3 1 /&**»"
             Adjusted, resisters  <^I3  asrt. R\K on
                              recaJULbrnXtd.
        3.  Results of audit following corrective action.  (Use format of  A, B, or C,
            above, as applicable. )
 II.  Calibration Drift Assessnent.
     A. Out-of-control periods.
            a.   Date(s)  7/fo /fS - 1 /3A /Sf       1/31- 1/5\/S5'
            b.   Number of days    4 _ •
      B. Corrective action taken 'ftg-ge.rQg^ gyrf rgQX^t^rztS^. On
                 -stlLI  drif^    because of-
                                                                       77?
Operator Signature            Date         Supervisor Signature            '  D
                                                                           ate
  To be  completed by the  Agency
                      Figure  9.3.   (continued).

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 18
    Most problems inherent to extractive analyzers will be electronic
in nature.  Procedures recommended  in  the  maintenance  manuals  can be
used  to  troubleshoot  circuit  boards  and components.    Extractive
analyzers are  often compact  enough so  that  in  the  case of  severe
electronic problems, they  can be  returned to the  instrument  vendor
for repair.

    The system operator will often be warned of problems by a loss of
signal,   inconsistent   readings,  or  poor   calibration  response.
Approaches to  resolving  problems  largely depend on the  skill  of  the
operator; however, some general guidelines can be given:

    •  Loss   of  signal  or   abnormally   low  values   -  check
       conditioning  system  for  plugging,  leaks, pump failure.

    •  Noisy,  erratic  signals  -  check  for electronic  problems,
       electrical   supply  problems,   weak   lamps,   moisture
       condensation,  particulate matter  in  analyzer.

    •  Loss  of linearity -  check  for sample  cell contamination,
       leaking  calibration  manifold,   incorrect  gas   cylinder
       values.

    •  Slow  response -  check  for  leaks,  water  in lines,  measuring
       cell  failures.

Failure  of  the  conditioning  system  and   consequent  drawing   of
unconditioned  sample  gas  into  the analyzer  is  one   of  the  worst
situations that can occur with an extractive monitoring system.  Once
unconditioned gas  enters the system,  it  may  take months  before  the
system will again operate properly.   Therefore,  redundancy and fail-
safe devices should be  built into the system from the  start.

    Table 9.2  lists some common problems  that  occur  with extractive
monitoring systems.   Those listed  range from the  physical  problems
often  associated  with  the  conditioning system  to those  associated
with the  analyzers.   Many of  these  problems  are due  to poor  system
design  resulting   from  a  failure  on  the  part  of   the  system
manufacturer  to  understand  the  constraints  imposed   by  the  plant
environment and stack gas conditions.  These problems  also may be  the
result of inadequate specifications  provided  by  the user at  the time
of purchase.   It  is difficult to foresee  problems and  it may often
require a redesign  of  the  system before the  frequency  of corrective
maintenance is minimized.

-------
                                                             Section No.  3.0.9
                                                                    Date  6/1/86
                                                                        Page 19
                 TABLE 9.2.   EXTRACTIVE SYSTEM PROBLEMS
    Common Physical Problems
     Possible Corrective  Actions
A.  Conditioning Systems
Probe plugging
Probe/filter  corrosion
Probe breakage (due  to  vibration or
   embrittlement  from chloride)

Condensation in sample  lines
Inadequate water removal
Dirt in sample lines,  plugged valves,
   plugged sample lines
Leaks in sample lines/fittings
Pump failure
Install blowback  system,  increase blow-
   back frequency probe shield.  Relocate.
   Change probe design.   Change system
   design.  Enter probe at downward angle.

Relocate probe.   Obtain corrosion-
   resistant alloy for probe construction.
Support probe.
   alloy.
Shorten.  Select resistant
Resize heaters.   Don't  let heat go off on
   heat trace.   Use  backup power.  Avoid
   shorts - don't loop  lines.  Install
   thermal conductivity sensor if
   continuing problem.   Remove water at
   stack probe.   Filter at lower
   temperature (acid may be condensing) -
   increase temperature or heat.

Improve chiller  design.   Connect two
   chillers in series.   Back up chiller
   with Permapure dryer (but heat front end
   of Permapure). Dilute the gas stream to
   lower moisture content.

Decrease pore size of probe filter.  In-
   crease sample flow rate to fine filter.
   Increase diameter of line.  Use clear
   Teflon tubing to  detect areas of
   accumulation.  Redesign to reduce number
   of valves.  Use redundant filters.

Reduce number of fittings as much as
   possible.  Detect leaks by pressurizing
   system and using  soap bubble indicator.
   Check for leaks in gas regulators.
   Don't wrench  down on compression
   fittings too  severely.  Don't use glue,
   paint, glyptal, etc.,  to cover leaks -
   rebuild system if necessary.

Perform routine  maintenance - check brushes
   periodically.  Check diaphragms of
   diaphragm pumps.
                                                                      (continued)

-------
                                                             Section No.  3.0.9
                                                                    Date  6/1/86
                                                                        Page  20
                          TABLE 9.2.   (continued)
    Common Physical Problems
     Possible Corrective Actions
B. Analyzers

Internal corrosion/damage
Poor response time
   (False positive  zero  values or
   poor calibration check  values)
Excessive drift
Component failures
   Lamps, fan,  chopper  motors

Electronic problems

   Loose circuit boards,  poor
      contacts
   Ground loops and noise

   Large voltage drops  when  plant
      equipment is  started.  Spikes
      in strip chart record.
   Static electrical  charges


   Burned out electronic  circuits
      from lightning  strikes
Check moisture removal  system  for  failure.
   Build redundancy in  system.   Add  extra
   chiller.   Put thermal  conductivity
   sensor in line to stop pump  when
   moisture  breaks through.  When  moisture
   breaks through, dismantle sample  cell,
   clean, and dry.  May have to replace
   entire cell in some  models.   Clean and
   dry all sample lines.

Check sample line length.  Shorten line or
   increase  flow rate.  Some analyzers have
   slow response times.   Increase  time for
   calibration gas flow during  daily
   checks.

Check fouling of sample cell for dirt or
   water.  Electrical problems.  Passiva-
   tion of cell surfaces.  Lamp weakening -
   light levels too low.   Detector prob-
   lems.  Electronic problems.   Erratic
   power supply.
Check component wear
   regular schedule.
Check and replace on
Check for vibration problems.   Install
   circuit board clamps.   Check  for  SO2
   corrosion in exposed units

Trace and rewire.

Install transient suppressor,  dedicated
   power transformer or constant
   voltage/isolation transformer for
   monitoring system.

Connect probe case to dedicated  earth
   ground.

Add phenolic gaskets between metal  stack
   and probe.  Add surge arrestors  at
   junction box.
                                (continued)

-------
                                                      Section No.  3.0.9
                                                            Date  6/1/86
                                                                Page 21
                        TABLE 9.2.   (continued)
    Common Physical Problems
     Possible Corrective Actions
 B. Analyzers

 Electronic problems (cont'd)

   No output from instrument,  no
      calibration cycle, etc.

   Improper instrument response -
      faulty calibration, improper or
      no output
Check fuses,
Check electronics.  Check  to see that cards
   and components are secure.  Use trouble-
   shooting guide supplied by vendor to
   check electronic test points. Replace
   appropriate components  or replace cards.
   Check software for errors in programming
   - particularly in calibration adjustment
   routines.
9.2.3  In-situ Gas Monitoring Systems

9.2.3.1    In-situ  System  Design.    In-situ  gas  analyzers  measure
pollutant  and  diluent gases  as they exist  in the  stack or  the  flue.
There are  two basic  types of  in-situ monitoring systems:   path  and
point.

    Path  monitors  generally  consist of  two  units,  placed opposite
each other across a  duct  or stack.   The systems use electro-optical
techniques  in  which  either  infrared or  ultraviolet  light  is  beamed
through the  flue gas.   Absorption of  light energy  at specific  wave-
lengths provides  a  quantitative  measurement  of  different  molecular
species.   Such instruments  can be designed to pass  the light  either
once or twice  through the gas.  The  once-through systems are known as
single-pass  monitors  and  the  twice-through  systems  as  double-pass
monitors.

    In single-pass monitors, the stack units consist of  a  transmitter
and a receiver.   The  transmitter contains an infrared or UV  lamp that
beams light  to the receiver unit directly across from it on  the  other-
side of  the stack.   The  receiver unit  senses the  transmitted  light
energy and analyzes  it to provide an indication of the gas concentra-
tion.  The transmitter and receiver  units are protected  from the flue
gases by  windows,  over which  a curtain  of  air  is  blown.   The  air
prevents particulate  matter from soiling  the windows, cools  the  parts
of  the  unit mounted  on the  stack,   and  prevents  the condensation  of
corrosive  materials  on the cooler  instrument windows.   The  purge  air

-------
                                                    Section No.  3.0.9
                                                           Date  6/1/86
                                                              Page  22
is generally provided  by blowers that draw  in filtered ambient  air.
The use of plant air for this purpose is not advised.

    Double-pass  in-situ  path monitors consist  of  a transceiver  unit
and a retroreflector unit.   Light  is transmitted and received by  the
transceiver.  The transmitted light  is bounced  off  the  retroreflector
located on the opposite  side  of  the  flue and returns to be analyzed.
The light passes over  a path  twice  the  length of  that of a  single-
pass  unit/   and  in so  doing,  may  be   considerably   weakened   upon
returning to the  transceiver  unit.   For  this reason, most commercial
path in-situ gas monitors are single-pass designs rather than  double-
pass.    Double-pass in-situ  monitors also use purge air  systems  to
protect window interfaces.

    Path  in-situ monitor systems  come  with a  number  of  accessories
that are  needed to protect  them from the often  hostile  environment
encountered  at  the  installation site.   For  stack-mounted systems,
protective hoods and covers are necessary to protect the transmitter/
transceiver,receiver/reflector units from rain, wind, and  temperature
fluctuations.   Lightning protection is  frequently  necessary.   Also,
anti-vibration systems are often required  to prevent the  optical  and
electrical components from shaking  loose.   A constant  voltage trans-
former dedicated to the monitor is also  frequently  necessary to avoid
large plant  voltage transients  from  affecting the  sensitive elec-
tronics of the transceiver units.

    Frequently, pipes are used  to  maintain the alignment  between  the
two cross-stack instrument units.  A modification of the support  pipe
provides a system in which an added outer pipe  can  rotate  to shut  out
flue gas  from  entering the  light  path.   Ambient air can  be used  to
purge out any  remaining  flue gas in the closed tubes  so that a  zero
reference reading  can  be obtained  by  the monitor.   Calibration  gas
cells are generally slipped  in the  light path  at this  time to obtain
an upscale calibration reading.

    Point  monitors  measure  the  flue  gas  over   a  small  distance
relative to the  larger duct  or  stack diameter.  This distance may  be
less than  5 cm  or, in  some  cases,  on  the  order  of  1/3  to  1  m  in
length.    The length of  the  probes  available   for  these systems  are
generally fixed, projecting the measurement section of  the probe  from
1 m to 2 or 3m into the stack.

    If  the  pollutant gases  are  greatly stratified,  the fixed,   one-
point measurement  may  not adequately  represent pollutant emissions.
The same  problem,  of course,  exists for the  single-point probes  of
extractive systems.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 23
    In-situ point monitors normally  use  a  ceramic thimble to protect
the measuring cavity  from particulate matter.   Pollutant or diluent
gases  diffuse  through  the  thimble,  which  excludes  the particles.
Because thimbles  may  become plugged  or  glazed over  in  wet scrubber
applications,  they must  periodically be checked  or  replaced.   Point
monitors that use ceramic thimbles do not require  blowers  as  do the
path  monitors,  but  still may  require  protective  hoods,  lightning
protection, and constant voltage transformers.

    Another type  of  in-situ  point monitor utilizes  an open probe to
support a  retroreflector.   Much  like  a  double-pass  path monitor,
light  from a  transceiver hits  the retroreflector and  returns  to be
measured.   The  measurement  path,  however, will  be on  the  order of
only a meter or less,  classifying  the instrument as an in-situ point
monitor, since  it measures  only a relatively small  distance  in the
flue gas.   In  this  case, purge air  blowers are  required to keep the
retroreflector and transceiver windows clean.

    In-situ monitors  provide  an  alternative to  extractive systems
since  they can  avoid  the requirements  for gas conditioning systems.
There  are  trade-offs,  however,  since the in-situ analyzers installed
directly on the stack may experience  severe environmental conditions.
Table  9.3  summarizes  some of the  advantages  and disadvantages of in-
situ and extractive systems.

9.2.3.2     In-situ   Analyzers  -  Absorption   Spectrometers  and
Electrocatalytic  Systems.   The basic principles  of operation  of in-
situ  systems  are similar   to  those  of  the  extractive analyzers.
Absorption spectroscopy and  electrocatalytic methods are  the two most
common techniques employed.

    The absorption spectroscopic  techniques  used in in-situ monitors
are

    •  differential absorption
    •  gas  filter  correlation
    •  second  derivative spectroscopy.

    The most  common  differential  absorption  systems are single-pass
path monitors  that  use a diffraction grating to distinguish between
measuring and reference wavelengths  in the UV  region of  the spectrum.
SO2 and NO are  measured by  this technique, although by changing the
optical system, it is possible to measure other  gases.   Filters can
also   be   used  to  distinguish   between  measuring  and  reference
wavelengths in in-situ  analyzers.   CO2 is  measured by infrared light
with this method.

-------
                                                             Section No.  3.0.9
                                                                    Date  6/1/86
                                                                        Page 24
        TABLE  9.3.   COMPARISON OP EXTRACTIVE AND  IN-SITD SYSTEMS
  Extractive Systems
                   In-Situ  Systems
                                    Path
                                        Point
Advantages
May be zeroed  and
calibrated with cylinder
gases

Analyzers can  easily  be
located in controlled
environments
Can time share  analyzer

Maintenance may not
require special training
                         Linearly averages  stack
                         concentrations  to  possibly
                         reduce stratification
                         effects

                         Does not alter  sample
Has fewer separate  compo-
nents than extractive
systems

Rapid response
                             Does not  alter  sample

                             May be zeroed and calibrated
                             with cylinder gases
Has fewer separate  compo-
nents than extractive
systems
Disadvantages
System (other  than
analyzer)  may  require
frequent maintenance

Probe plugging possible
in dirty gas streams
May alter sample


Long sample lines  reduce
response time

Zero and calibration
gases expensive
                         Can monitor at  only  one
                         location
                             Can monitor  at  only one
                             location
Difficult to repair  on  site   Difficult  to  repair on site
May lose light levels  in
dirty gas streams  or  in wet
scrubber applications

Requires temperature
compensation
Generally cannot  be  cali-
brated with cylinder gases.
Audits are expensive: must
perform reference method
testing

Special training  or  vendor
service often needed
May clog in dirty  gas
streams or in wet  scrubber
applications

Requires temperature compen-
sation
                                                      Special training  or vendor
                                                      service often  needed

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 25
    Gas   filter  correlation  spectroscopy  is   a  nondispersive
spectroscopic  technique  that  has  been  used  in  single-pass  in-situ
analyzers as well as in ambient air analyzers.  The method requires a
gas cell to be filled with the pollutant or diluent gas that is to be
measured.  Light, usually in the infrared, is transmitted through the
flue  gas  into  the  receiver unit,  where  it  passes through  the  gas
filter cell and  then  through a neutral  density  filter  (or  no filter
at all).  The  gas  filter  essentially "filters out" the spectroscopic
regions in  which the  pollutant molecule absorbs  light  energy.   This
100% filter thus gives a  reference  signal  to  which the light passing
through the  neutral  filter can be  compared.   The  advantage  of this
method,  as  well as  with  the  differential absorption  technique,  is
that   low   levels   of  particulate   matter   will  not   affect  the
measurement.   Gas  filter correlation  instruments  have  been designed
to measure SO2/ NO, C02,  and CO.

    Second  derivative spectroscopy  is  employed  in a  point  in-situ
monitor  that  has been quite  widely marketed.   Using  an oscillating
lens,  UV absorption  peaks  of S02  and NO are  scanned 45  times  per
second.  This  scanning creates a signal  that  is  related to the second
derivative of  the  absorption  peak,  taken with respect to wavelength.
Using  the Beer-Lambert  Law,   it  can  be shown  that  this  signal  is
proportional to  the  concentration of  the  pollutant  in  the  flue gas.
The  instrument has a measurement  cavity,  generally 5  cm in length,
that is protected by a ceramic thimble.  The  system can be calibrated
by using either gas cells or cylinder  gases.

    Electrocatalytic  analyzers  used  for   in-situ measurements  are
exclusively  point  monitors.    Currently,  this  technique  has  been
applied  for  the measurement of 02 and SO2.   As  discussed in Section
9.2.1.4  for  extractive  analyzers, solid electrolytes can be  used to
generate  a  measurable flow  of electrons.   Here,  a  ceramic  thimble
keeps  the  measurement  side  of   the  solid  electrolyte  free  of
particulate  matter.    Calibration  gases  can be  injected into  the
measurement cavity to check the instrument operation.

9.2.4  Recommended Maintenance - In-situ Monitoring Systems

9.2.4.1  Operation Checks  (Daily Checks).  The daily operation checks
associated  with  in-situ  gas  monitors  are   similar   to those  for
extractive  gas monitoring systems.   The  operation  checks  should be
performed  by  a  trained  and  qualified  operator  who has  been given
responsibility for the  system.   The  monitoring system  will  have a
better record of performance if the operator  checks the  system daily-
automatic zero and calibration procedures  can create a sense of false
confidence that  can lead to system failures.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page  26
    The  daily operation  checks  should  start with  a review  of the
previous 24  hours  of strip chart  recordings  and computer printouts.
Discrepancies  should be  noted  on  the  check  sheet  and instrument
logbook.  Again, any system faults identified by the  indicator lights
should be resolved immediately.

    Many in-situ systems use a gas cell containing pollutant gas at a
known  concentration for  calibration. In  the  optical  in-situ  ana-
lyzers,  the  cell is put  into the  path  of  the  light beam,  and the
signal is attenuated by a specific amount.  This method may not, how-
ever, always check the  complete optical system over which  the flue gas
is measured,  although  the  use of pipes that  close out the stack gas
from the light path  have, in some cases,  alleviated the problem.  Also,
the gas  cell pollutant concentrations are  not  normally  certified by
independent  laboratories (their  concentrations  may,  however, be  inde-
pendently verified).  Degradation of cell concentrations by leakage,
adsorption,  or  internal  reactions  can  cause  discrepancies  in the
calibration data.

    It is  possible  to  calibrate  some in-situ monitors with certified
cylinder  gas.   This  can  be  done  with  in-situ  point   monitors  by
flooding the  volume within the  ceramic thimble  with calibration gas
or with  zero gas.   A  "flow-through"  gas  cell can be used in either
single-pass or  double-pass monitors.   By flowing gas of a certified
concentration  through  a  fixed cell in the  instrument, a calibration
"traceable"  to  NBS or  other  certified    gases can  be  obtained.
However,  problems  associated with  the  optical  path  used   in the
calibration sequence may still be  present.    It  should also be  noted
if the temperature  compensation circuit  is  disconnected in  any of
these  calibration  sequences.    Problems  in  these  circuits  may  be
overlooked in such cases.

    The  optical alignment of the components  of in-situ  systems is
very important for proper operation.  Alignments should be checked at
appropriate  intervals.  Many  of  the systems  do  not, however, incor-
porate alignment sights in  their  design.   In  such cases,  detailed
electronic or optical checks may need to be performed  to optimize the
system alignment.

    Note  again  that  actual  calibration  adjustments  should  not  be
performed  arbitrarily.    Control   limits   should be set  to   avoid
calibrating   against  normal   system  fluctuations.    The  Federal
monitoring requirements specify  that that the calibration be adjusted
whenever the  zero  (or  low-level)  or  the  high-level  calibration  value
exceeds two  times the  limit  of the applicable performance specifica-
tions. *  Performance specifications  are given in Appendix B of Part
60 of  the U.S.  Code of  Federal  Regulations  -  Title  40.   If either

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 27
calibration  value  exceeds   four   times  the  applicable  performance
specification during  any calibration  drift check,  all  of  the data
collected since  the last check are  considered to be  data collected
during  a  CEM   system  out-of-control  period  and   are   rejected.
Similarly, if  the system exhibits  drift  (either  low-level  or high-
level) greater  than  the  performance  specification  for five  conse-
cutive 24-h  periods,  the  system  is again  viewed as  being  out  of
control,  and  subsequent  data are  rejected  until  the  system problems
are corrected.5

    Figure 9.4 gives  a format for developing a  daily  QA check sheet
for in-situ systems.

9.2.4.2   Routine Maintenance  (30-day  Checks).   Routine maintenance
should be initially  performed on  in-situ monitors,  at  a  minimum,
every  30  days.   Although many  equipment  vendors  recommend  routine
maintenance at  periods of two or three months,  it is  advisable  to
gain experience with the system before waiting these longer periods.

    Routine  maintenance  for  in-situ  systems   should  consist  of
ensuring that key components of the system are clean and operational.
In general, windows, filters, and desiccants should be cleaned and/or
replaced.

    In  the cleaning  procedures   for  windows or  optics,  great care
should be taken.  Lens tissue or clean, soft rags should be used with
a solution of alcohol and water.    In severe cases, mild detergent may
be  necessary  to  clean  windows exposed to  the  flue gas.   Sensitive
optical  components  such  as  diffraction  gratings  should never  be
touched or cleaned  in  the field.   Fingerprints or traces of cleaning
materials  can   severely  affect  their   performance,  so   special
techniques must be used.

    A  systematic  procedure  should  be instituted  for  checking  the
electrical performance to compare it  with  the  original  factory  or
start-up performance.   This  normally involves using a digital volt-
meter and oscilloscope  to check  the  analyzer  at  various test  points.
These  test  points  should  include  a check  of  lamp voltages, power
supply voltages,  and detector outputs.   The procedures  involved  in
these  checks  will  generally  require  a  well-trained   serviceman  or
electronic technician.

    Many  in-situ  analyzers  use  chopper  motors  in the   transceiver
assembly  to modulate  the light beam or  switch instrument  functions.
These  motors  should be  checked  for bearing  noise or  for excessive
vibration.  For  motors that automatically move  mirrors  or gas cells

-------
                                                   Section No. 3.0.9
                                                         Date 6/1/86
                                                             Page 28
Example Format for In-sLtu Gas Monitoring System
Daily QC Checks
Plan: C$L\ Vfljrtfgf / kCjfrle. Date 1/28/85 Time \O\\loCLJT\
Unit fipULe/v *•' C/UtipJ"
Gas Monitsred 5fc>j /C.OA
*nal/2er 1.0. CrOSiS^T-^rV"! 1— A5O
Scan Value 1 OoO rarn -^Pp/IE* to.
Calibration Value BYa. ia%C^>
icxiir.-ier cis, -;js ceUil
Zero Gas (air, N?, other) W/4

Part 1 Indicators
Indicator L-Tr.ta S:a
Mam. -RoteH= Exwrnee^
Phone fiq-5t-q-.au.
Offset Value /O j^pno

So*.
Paoer Status:
Str ip chart OK
Printer ^l^

•us 1 Problem/Ac ti an taken
Power failure Q^
Blower failure ; ...
Lamp failure OK j
0lcty "ind" at (u-^artKO Mote, te ol«n^ new:t tuett
Alignment Q^.
*lacrn UluVlu'n U»t«>«
Oth'r i . /Vo«'.5&
Gas cell 12 || ; 36
Calibration qas
{ if aoclicabU)
'


,

a 1 13. pp^vt i? a.
| ^ia pp^*i "^sio ^oa-
i
! IA. ^ la.o i A. j
i
i 385 pf»vi 31^ 37S

Part 3 Zero and Span Adjustment (if outside of control Liaits)
Zerocontrollimitt 5 Don

Ad jus fed Read ini a | Tin*
z«f<> i ia;i5 p<
Gas cell » I , 1^-. 3^

Meter ! 5:.-.? Cr.art Digital Printer
^ 0 ! 0 0
\ iia. • lia 913-
Gas cell  ll-O
Calibration ^a? ! - I
5tac'< ::r;c »-.- : i- icr. , : :
a==:ic.«:*; i:iti
Hft>tYt Cl Vt I V<^|- %1 /.M"/
Operator ai-qr.dt-rs "^'e
! 3^3. : 3U-3. 3H-3-
at 	 4^_ jU,\ir
„,.„.,„ .^ N ,,

Figure 9.4.  Example format for in-situ gas monitoring system daily
                            QC checks.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 29
into position, the operations should be  manually  checked to see that
the movements are smooth and complete.

    Observations made during the routine maintenance period should be
noted in  the system  logbook  to provide  the  base data  necessary to
optimize the maintenance procedures.  After six months to one year of
operation,  trends  in performance  should  become apparent.    These
trends may  indicate  that  some maintenance procedures may  need to be
improved or  that  others may need to be  performed more  frequently or
less frequently.

    Figure 9.5 gives a suggested format for the development of an in-
situ system maintenance check sheet.

9.2.4.3   Performance  Audits.   The  discussion  in  Section 9.2.2.3 for
performance  audits  on extractive systems  is  also applicable  to in-
situ  systems.   However,  in the case  of path  in-situ  monitors that
come  under  the  requirements  of Appendix  P,   it  is  not  possible to
conduct a cylinder  gas  audit  (CGA)  unless  a flow-through  cell  is
incorporated in the system.   RATA  and  RAA audits can, however, still
be conducted on these systems.  Path monitors should be checked quite
frequently  to  see if they are giving  representative  data,  since, as
mentioned  previously,  the  methods   used  for  internal  zero  and
calibration  may  not  check the complete optical system.   The  in-situ
point monitors can,  in most cases,  be  checked with cylinder gas, but
audits independent of the monitoring  system should also be performed
at regular intervals.

9.2.4.4   Corrective  Maintenance  (Problems and Troubleshooting).  In-
situ  monitors  avoid  many  of  the problems associated with  extractive
systems;  however,  in-situ systems  are  not  devoid of problems.   In-
situ  systems  installed  directly  on a  stack or  duct  are subjected to
severe environments.  Temperature  cycling,  acid gases,  and vibration
can  damage  the  sensitive optical  components  and alter  the  optical
alignment  that  is  important  to   these  systems.    The   electrical
components  located in the mounted  system assemblies  are subjected to
the same type of abuse and can fail if not adequately protected.

    The often complex  nature of  in-situ systems requires  a higher
level of  troubleshooting  skill.   It  is  advisable that  the operator
responsible  for  a  CEM system ' first  be trained by  the  instrument
vendor, preferably  in  a  formal  training  session  at   the vendor's
facility.

    The basic  problems  that occur  are  often a result of the environ-
ment,  such  as vibration  and ambient temperature  variation.   The
solutions to these problems are very specific to  a given  installation

-------
                                                   Section No.  3.0.9
                                                         Date 6/1/86
                                                             Page  30
Example Format Cor In-situ Monitor
30-day Maintenance Check Sheet
Plant CcL.\vafY\0C /Acsr\€. fhwef Date H/Sk./?^) Time /0;DS QJVI
unit TboiLp/i Wo- i - Ou*i»i: Name "Robert £rcu*ef>r
1.T
••oni'.ot 1.0. Ho. CjrCf¥Jtf>cb\ L2.50 Pr.one 1 lirtj/
& (jDmtenaxJwi.
—
Vi'3uo.liy dlir-ty
Claecktct OC.
Action
CoHec*€d/ ne«io«^C*
t Icjcaji ct^d>ve^
lUipedi el«tr\
Ouiperf cteouxv
LUip«d* ci-eo-i^.
CX->rf«* cJeox\

(Wifxa* cJwa*v_
OK.
OK
Clax^ei ptvsefRA£tito\
CWeoLned
C.i«ui«e)
"Ref3tac«d5
I\JO OLtVtiorv

Status
OK
g^Biue n Red
C>K
OK

RcLsjx^g
Ac t ion
—

—
-

"R«pU«jr,«GL brushes S/ia-/J5
"Robei-t SrviLoepr 4/'S>-/8-5 CAc — ^ 6u
-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 31
and  may  require   re-installation  or  the  construction  of  environ-
mentally controlled enclosures around the analyzers.

    Table  9.4 lists  some of  the problems  associated  with in-situ
systems and recommendations for corrective action.

9.3  Opacity CEM System Operation Practices

9.3.1   System Design  -  Many  types of instruments have been  developed
that can  measure  the opacity  of a  gas  in a  stack or  duct.   These
instruments  are  called  opacity   monitors,   or  alternatively,
transmissometers.    They  are  in-situ  monitors  and can  be  designed
either as single-pass or as double-pass units.

9.3.1.1  Single-pass Systems.  Single-pass opacity monitors  are quite
simple  in design.   Light emitted from a lamp passes through  the stack
and is  sensed by a detector  on the other  side of the stack.   Blowers
located on each side help keep the windows of the apparatus  clean.

    One  problem  associated with  the design is  that  of calibration.
The electronic and optical components of the system cannot be checked
unless  the  stack or duct  is  free of  particulate matter.   This can be
achieved  if the  industrial process  is  shut down  and provisions are
made so that  residual dust does  not  recirculate in the stack.  Since
most industrial processes cannot shut down just  to  zero and  calibrate
an  opacity monitor,  instrument  designers have  used  optical  light
fibers  or  zero-pipes  to  provide  this capability.  The  light fibers
pass from  the  lamp to the detector  around  the  outside of the stack.
The zero-pipes pass  through  the stack and can  be purged with air to
provide a zero reading for the instrument.

9.3.1.2  Double-pass Systems.  In double-pass opacity monitors, light
crosses  the stack and  is returned  by  using a  retroreflector.   The
retroreflector  returns   it  to  the  main  analyzer  housing,   where  a
detector  then  senses  the light.   This  instrument  can  be  zeroed and
calibrated by flipping a mirror up into the light path to approximate
a  clean-stack  condition.    The  mirror   close  to  the  transceiver
prevents the beam  from  crossing  the  stack and merely sends  the light
through  the  clean interior  to  give a  "pseudo-zero"  reading.    A
calibration filter can then be flipped into the  path of light to  give
an  upscale calibration  reading.   Rotating  choppers   have  also  been
used for zero and calibration procedures.

    Blowers again are used to  keep clean  the optical  surfaces exposed
to the  stack gas.  Filters are needed before the  blower so that clean
air will pass through the analyzer.

-------
                                                             Section No.  3.0.9
                                                                    Date 6/1/86
                                                                         Page  32
                TABLE 9.4.   IN-SITU GAS ANALYZER PROBLEMS
              Problem
      Possible  Corrective  Action
Excessive dirt buildup on windows  or
   thimbles
Cyclic drift in signal  unrelated  to
   plant performance -  due to  ambient
   temperature changes.   Signal
   becomes erratic from high
   temperatures

Optics misalignment/electrical noise
   due to stack or duct
Signal becomes erratic  at  high
   opacities
Misaligned system

Probe/seal leaks (in-stack  monitors)

Lightning strikes
Static charge buildup burning  out
   circuit boards

Lamp burnout/degradation
Gas cells unreliable


Spurious readings during plant
   start-up,  shutdown,  etc.

Improper temperature compensation


Increased response time
Clean blower filters on path  systems.   In-
   crease blower capacity.  Rotate  ceramic
   thimbles on point systems  or  replace.

Insulate protective hood or install
   temperature conditioning system  about
   monitor.
Mount assemblies independently  from  duct.
   Use flexible bellows  for  duct
   connection.   Dampen mountings.  Relocate
   monitor.

Relocate monitor after control  device.
   Water droplets from scrubber may  also
   cause this problem.  In  such cases,
   analyzer  may have to  be  located before
   the scrubber.

Realign, check, and tighten  system.

Return to vendor for repair.

Add phenolic gaskets.  Add  surge  arresters.
   Relocate  monitor.

Run copper cables to earth  ground.


Replace on regular schedule.  Check  lamp
   power supply for high incidence of  lamp
   failure.

Cell leakage, losses to  walls - recalibrate
   cells or  replace.

Install transient suppressor, dedicated
   transformer  for monitoring system.
Adjust circuits,  recalibrate,
   or replace boards.
or repair
For in-stack in-situ monitors,  check
   ceramic thimble for  plugging.   Replace
   if necessary.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page  33
    Any transmissometer, be  it  single-pass or double-pass, will have
similar components, including a light source, a detector, and various
optical components.   The light  source may  be  specifically selected
for the system or may merely be a lamp normally used in an automobile
headlight assembly.  Detectors are chosen so that they will sense  the
light emitted by  the  source.  This  light  is normally  in the visible
and  infrared  regions of  the spectrum.   Optical  components include
focusing lenses,  mirrors, and optical filters.   Filters  are used  to
help the  instrument  respond only to light  in the  visible region  of
the spectrum - the so-called photopic  region,  to  which the human  eye
is  sensitive.   Filters  normally prevent  infrared light  from being
sensed by the detector.

    Mirrors, of course,  direct  the  light  from point  to point  inside
the  instrument, so that  the  operations  of  measurement,  zeroing,  and
calibrating can take  place.   Lenses focus  the light  so that a well-
collimated beam is sent  across  the  stack and through the  instrument.
In addition to these  components,  chopper wheels  and light modulating
wheels are often used to direct or modify the light beam.

    Various  accessories  can be purchased  for the  opacity monitor.
These  include  integrators,  beam combiners,  "reasons"  panels,  and  so
on.  The main  instrument  panel  for  the monitor is normally  installed
in  the  control  room  of  the  plant.   Here,  the instrument outputs  are
observed and recorded.

     Transmissometers are generally  installed to meet requirements  of
pollution  control agencies.   The  Federal  EPA has  developed  speci-
fications  for  these  instruments.1   There are  basically  two types  of
specifications:    design  specifications  and  performance specifica-
tions.  The design specifications detail how the instrument is to  be
constructed.

    Monitors that  satisfy the design specifications are purchased  and
installed  in a  location  according  to EPA guidelines.1  The  installed
monitor  undergoes  the   performance  specification test   procedures.
These procedures check the system for  zero  and calibration drift over
a  one-week  period.   Passing  this test  indicates  that  the monitor  is
acceptable for the opacity monitoring.

    However,  testing should not  stop  at  this point.    As  with  all
instruments,  problems  can   occur  over  long  periods   of operation.
Blower  filters must  be  checked  and  regularly  cleaned, burned-out
lamps  replaced,  and  the  integrity  of the  data  must  be  consistently
audited.    The  continued  success  of  a  monitoring  program depends
heavily on how well the instruments  are maintained.

-------
                                                     Section No. 3.0.9
                                                           Date 6/1/86
                                                               Page 34
9.3.2  Recommended Maintenance  -  Opacity Monitoring Systems

9.3.2.1  Operation Checks  (Daily  Checks).   Operation checks should be
conducted  daily.    These  checks  can  generally be  performed  at  the
remote control  unit  located in the  station control room.   It is not
necessary  to  go into the  plant to check the  transmissometer itself,
unless a fault  lamp indicates a problem has developed.

    At this  level, the  activities include  noting  the status of the
system indicator  (fault) lights and  recording  the  daily zero and span
values on  an  appropriate form.   The day's strip chart record should
be  examined  to  check   for  trends  or  problems  that  might  not  be
identified by the system fault  lights.   At  this time,  the strip chart
should be annotated.    Figure  9.6  gives  examples  of  the  type  of
information that should  be noted.
1OVU
1700
1600
1500
1400
1 1300
1200
1100
1000
onn










C
















*
^
i
I








i
^|
^
e,
^
-i
i
j


•
=3
^l
/
^
^i
s5
1,
IP
a —



























1:15 am





















4/
"~Ch
FS
Of






















































20/84
art speed 5 cm/h
= 100% Opacity
fset=10%





















r —



                    -10  0  10 20 30 40 50 60 70  80 90 100
                           Average opacity, six minutes

                       Figure 9.6.  Strip chart
                             annotations.
    The  first  level of quality  control serves to  alert  the operator
to  problems  or  necessary  adjustments.   If  the  window  indicator or

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 35
zero compensation light (where applicable) warns that the windows are
dirty, someone  will  have to  go  to the  duct  or stack  and  clean the
windows of  the  transceiver  and  retroreflector  assemblies.   If the
span  reading  or  zero reading has drifted  to  an unacceptable degree/
the span or zero potentiometers may have to be adjusted to bring the
readings back to the proper values, or other corrective action may  be
necessary.  Care should be  taken in  such  adjustments,  however.   If
there  is  only  a  small  difference   from the  proper  values,  the
difference may be due  just  to random noise in  the  system.   Also,  if
the drift is  large,  problems  may  be  developing that cannot be solved
by merely adjusting  zero and span potentiometers.  For  these reasons,
"control limits" are commonly  established  to  aid in deciding whether
to adjust the system.  The specific limits  should be established at a
level  that   (1)  ensures  the  monitoring data  will  be sufficiently
precise for its intended use and  (2) is  achievable.

    To make  the daily  check  thorough  and complete,  log  sheets are
often developed by the operator.   Figure 9.7  should be regarded as a
starting point for developing such a form.

    Part  1  of  Figure  9.7  provides  a  checklist  for  the  indicator
lights on  the remote  control panel.    Items  incorporated  in  Part  1
will  depend  on  the  specific  instrument system.   Depending  on the
status of the lights, the daily operation check may develop into more
than  just  a  check.    A blower  failure or  dirty window  alarm will
require a visit to the transmissometer for  corrective action.

    Note also  that Parts  2  and 3  of Figure 9.7 require data from the
meter, the strip chart,  and the  digital printer.   If  the  monitoring
system presents data in all three of these  formats, the data obtained
from  each should be  identical.  If they  are not, the correct one must
be  determined.   In poorly  maintained  systems,   improperly  zeroed
meters or  recorders  can create discrepancies.   Also, meter readings
often differ  from the computer printout because the  transmissometer
and computer are improperly connected.

    Zero compensation values should also be recorded so that they can
be evaluated at a later time.  After several weeks, these data can  be
drawn from the daily logs to evaluate the rate of window soiling.

    As a  part of the  quality assurance program,  the  form should  be
signed  by  the  person  performing   the  checks.     At  appropriate
intervals, the  operator's supervisor  should  review  and  initial the
logs  to see that assigned responsibilities  are being carried out.

9.3.2.2   Routine Maintenance  (30-day  Checks).   The  second level  of
quality control  for  transmissometer systems  involves establishing a

-------
                                                    Section No.  3.0.9
                                                          Date  6/1/86
                                                              Page  36
Example Format for Transmissometer Daily QC Check Sheet
Plant LaJyoyY+er /Acvvre ffcujer Date ^/SiS'/KfT Time /'. /5"pK>n
Unit uQilfff No. 1 OtrOe-t Name 'Birjljpr't. Eri«u-ieer
Transmissometer 1.0. Ho. CrDSsfecH Tl.ll Phone fl<^- SfR-IS^M-
Span filter value ?R.p-^ Stack-exit correlation value (l*/lr) Q.9/
Hours ooeratinq in period: ^'f/3'V Paoer status: strip char; <3K
printer OK

P»rt 1 Indicator Lights | Status Problen/Accion taken
Power failure o^
Blower failure ....
L?(S
Lamp failure olKi«.
Other _ j

Meter Strip char; {Digital printer
Part 2 unadjusted Readings | Ti:"e * °P ° ' °? | D j < 0? | D
2eco iiaopm 0.6 o.ooa. e>. a. | o.eoi 1 O-3 [o.oot
sp«n r.a6 at.5 o.iaf AU.I (0.131 | at.j. jo./sa.
Stack opacity ,130 Jv.S' O.07? ' /fe . / 1 O.O7fe 1 /(p. A. |<5. 077
Zero compensation 1' 36 I.A. 0 005
Part 3 Adjusted Zero and Span Readings (if outside of control Units)
Zero control limit » 	 f£ ^ Op, Span control limit « 	 £ * ^P
Meter Strip chart Digital printer
Time 4 Op o A Op D J Op D
Zero 1:35 0 f
Zero compensation ^ . qg / ^ O.OOS' windows cleaned? Yes Q *° &
Comments/observations:
'Rc'jjgf^j S^ci^tear" l/J-f/SS^ fJU*— <^\ OAX\.^S <^\x'^\li5')
Operator Signature Date Supervisor S^igr.atiUaJ Date






























Figure 9.7.  Example format for transmissometer daily QC check sheet.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 37
schedule for  routine  maintenance.   A basic  schedule  may be provided
in the instruction manual, but such schedules do not address problems
specific  to  an  installation.   Over  a  period  of  time,  after  the
operator  is   comfortable  and  experienced  with  the  system,  a
maintenance routine can  be  developed.   It may take a  year or two to
perfect the schedule,  but once developed, it should be adhered to.

    It is recommended  that,  at  least  initially,  a 30-day  maintenance
routine be  established.   Housekeeping  of  the  transmissometer equip-
ment located on the stack is  the predominant form of maintenance per-
formed at this interval.  Cleaning of outer  surfaces, windows, parts,
and filters is  the  major activity.  Once  the  operator or  serviceman
is  at  the  instrument site,  the  alignment, desiccants,  and blower
motors can also be checked.

    After a few  system failures  have  been experienced, a  replacement
schedule can  be  developed for parts having  a  lifetime  of a year or
more.    For   example,  transmissometer   lamps will  generally  have  a
lifetime of over two years  (20,000 h).  Rather than allowing the lamp
to burn out and  cause  a  system  failure, it  can be replaced  routinely
at  the end of  every  two-year  period.    Blower  motors  also  require
maintenance.   Since  the motors  are "on"  continuously,  at some point
the bearings  are going  to wear  out.   The period  of time  it  takes for
this to happen will depend  on several  things - such as the  amount of
particulate matter  and  acid in the  ambient air  and the amount of
stack  vibration.   These  factors  will  aggravate  wear  to  cause  a
problem eventually.   When breakdown does occur,  the period  for motor
maintenance can then be  established.

    In the  transmissometer  check sheet  given in Figure 9.8,  there are
two  parts,   one   for   required  maintenance  actions,   the  other  for
required  maintenance  checks.   The  serviceman  or  operator  should
perform the basic housekeeping  chores  of cleaning and then  check for
other  problems.   If  a  problem  is observed  or  if one  appears to be
developing, corrective action should then be taken.

    The  form  should  be  modified  to  fit both the  plant  maintenance
routines and  the characteristics of the actual system.  Within a year
after  the system has passed  Performance Specification 1,  the operator
should  use  the   logbook and  daily  logs to  devise  a  schedule  of
periodic preventive maintenance that best fits the installation.

9.3.2.3   Performance  Audits.  The  third level  of activity  that a
company should  institute for its  opacity monitoring  system  involves
conducting performance audits.

-------
                                                   Section No.  3.0.9
                                                         Date 6/1/86
                                                             Page 38
Example Format Cor Transmissometer 30-day Maintenance Sheet
Plant (la.\ypLr^e.r /Actvie fewer Date M/ia./S'Sj Time /( '• \S ajr\
unit Boiler- Wo. 1 - Outlet Name "Robert Sncti^eer"
Transmissometer I.D. Mo. CrPSStech "Tl^l
Part 1 Required Maintenance Actions
Clean outer surfaces
Hood
Transceiver assembly
Retroref lector assembly
*
Purge air system
Clean inner surfaces
Hood
Transceiver assembly t'jnlatchedl
Retroref lector assembly i^-j latched)
Preseparator - air purging system
Clean transceiver windows
Clean retroref lector windows
Clean fiber-optic cables (if applicabls)
Replace purge air filter (or clean and
replace)
Tighten hose cla.Ttps, cables, mountings
Phone • £T;<1'-5HC1- ISI.M-


Status

Dosl-
J>ir^
Di'rt^

D,r-t^

PLUn
OK
OK
Diri^
D/rgT^' '
^';Ht
NA
92-%

Part 2 Required Maintenance Checks
Status
Check alignment (?) Misaligned
Check desiccants Blue (3 Red LJ
Check cables - continuity, pinching/ T/Txx>=c«iver coJOK-
cuts, corrosion leoje
Check hoses - continuity, pinching/
cuts, corrosion OK
Check security seals 0K.
Blower motor - bearing noise £,^
nob^ydl" £wci/T>eer' H/iays?> Cx/i-.^c^
Action

^rlS^J-eo^ed
U^iped1
Luif>^»
UJ.perf

W.ped
(\JOioe
(\^sne.
Cu.pcd
C ^Ga.^&A
CI«L.r»:C*

7.ep^^


Ac t ion
(V) Realigned

"^•^ ^^
-
-
_
6" ^Ml<'

Operator Signature Date , Supervisor Sjignatu^je/ Date
Figure 9.8.  Example format for transmissometer 30-day maintenance
                           check sheet.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 39
    Audit  devices  have been  developed for  most of  the double-pass
transmissometer  systems  and  for  some  of  the  single-pass  systems.
Basically,  the  devices designed  for  double-pass systems  consist of
holders  that  can  be  attached  onto  the  transceiver.   The  holder
consists of a  slot  for holding calibration filters  and a short-range
retroreflector.  The device and the transceiver basically constitute
a  "mini-transmissometer"   that  can  accommodate  audit  calibration
filters.  The  device also  contains an iris,  which allows the reflec-
tance of the  mirror to be adjusted so that it  will  correspond  to a
simulated,  stack zero.  It is  best to make this adjustment initially
when  the  system  is  set  up  for   the   performance  specification
calibration  and  response  time  test.    After  the  instrument  is
installed,  it provides a method of checking the  simulated, instrument
zero.  This "audit zero," of course, may not necessarily be identical
to a clean-stack zero, but comparing  it to the  instrument's internal
"simulated zero" provides a good quality control check.

    An audit does not just involve  obtaining data from audit devices.
An audit is a  check  of the performance of the entire transmissometer
system.   Indicator  lamp status,  stack-exit  correlation corrections,
alignment,  and other functions of  the system should all be checked at
this time  by  the  auditor.   The  most  common problems  uncovered in
audits  are  errors   associated with  the  calculation of  stack-exit
corrected  opacity  values.   An auditor  can  easily  uncover  such a
problem through a well-designed program.

    A  system of auditing  procedures  has  been  developed for double-
pass transmissometers.6   This system, designed for use  by control
agency  personnel,  gives  detailed  step-by-step  instructions   for
conducting an  audit using the calibration filter audit  devices.   A
generalized  form,  based  on the work  detailed  in  reference 6,  is
presented in Figure 9.9.

    Figure 9.9 indicates the type  of  information that can be obtained
during  an  audit.     Using a   ^calibration  audit device,  both   the
transceiver optics and electronics are evaluated.  Using a reference
signal  source,  the  data  handling system  can  be evaluated.   In  the
case  of  a  double-pass   system,  these  procedures   check  only   the
transceiver  assembly and  data  handling  system.   Since  the  retro-
reflector assembly is not involved  in the  checks, the audit evaluates
only part  of  the system.   It  is possible that  a system can pass an
audit  without problems,  but.  the  cross-stack  opacity  readings   may
still be inaccurate  if misalignment or window fouling problems occur
at the retroreflector side of  the  instrument.

-------
                                                                                       Section  No.   3.0.9
                                                                                                  Date  6/1/86
                                                                                                         Page   40
                       Exaaple Format for Transmissometer  Performance  Audit  Data Sheet


                 Plane  CaJvorVW /Acwif.  ftxugr	    Date    l/'Wf5     Tine  ,
Unit   'BpJer  Mn.  I - Outtefc

Tr>n»iiioncir 1.0. »0
                                                                .Cro.
                                                TlJJ
                    A.  Stack-e«it correlation
                       1. Emission outlet pathLengtn, ln
                       2. Monitor pathlength, /,                          I	&_2_*e:a:i
                         :alculated ratio I*,


                    S.  ault indicator lamps                                OM      CFP

                         Power
                         Otrtf window
                         Air purge
                         Alarm



                 C.  Internal zero a.id span che
                       1. Internal span value
                       2. Internal span value 'crrr»cc»d  co JCJCJt exi-)         	   ^?-£	  *
                                  Opx«t-< l-'PuMnmrtfl'v't
                       3. Remote control unit ^etar readings (corrected)
                                  Zero raadina                             	     Q.C	  A
                                  Span read I .ig                                  ^"»       «
                       4. Scrip chart readings (corrected)
                                  Zaro raadinq                                   ^-^   —  ^
                                  Span reading                           ..—.    , ^i V	  i
                       9. Oiqttal printout icorrecced;
                                  Zero reading                           ,.—.,    , J^»^	  \
                                  Span reading                                 —*^- fc	  *
                                  Averaging time                           	     \£. GiS"	,3. C^  \ op
                                  Before cleaning                        _.
                                  After cleaning                                      r .<^_
                                      Time cleaned  Jt'C-5

                       2. Alignment chack-stacus                          —      ^	
                       3. Other mater performance  indicators 'if dpplicdM*.'          ^A
                                  Raf signal                                   £**
                                  AGC


                    (.  Recneck of instrument intarnaL calibrations
                                  Tine                                	//*'/6 L/J»/	
                                  2«ro reading                                  ^-^
                                  Span readi.-.g                          	3±LJc*	
                                  Zaro compansation                     	ii	

                 II. Calibration Device Audit
                    A.  Audit device lero
                       L. «ater    0.0   *0p
                         Strip char: Q.^L *Cp
                         Print output &Q    "^Op
                       Low-rang* filter
                         Filter number
                         Filter opicity (urccer*cte4)    /C. A
                         Filter opacity fcorr«ec*i>       9.. 3
Figure  9.9.
       Example  format  for  transmissometer   performance
                      audit  data  sheet.

-------
                              Section No.  3.0.9
                                    Date 6/1/86
                                        Page 41
II. (
c
[
III.
IV.
_J
alibracioo Device Audit (Continued)
I. Filter nuaiber A£~_/*fF
2. filter opacity (uncorreccedj 30. ¥
3. filter opacity ('corrected^ /#, X
t. High-range filter
2. Filter pacity funccr recced/1 •*//. 3.
3 . filter pac i t y f corrected 1 3S-3

finte^r ted corrected opacity values frost systeta cuepuc - computer printout
or eiyui alenc)


| Tiaie i | Time j T:=e j

I ! r.Cfp~.\ //.f i l;si> 1C.7 1 /•>>- '
2 . /.•« j ii? • ! a: /a. 30.1 '. W •
1 i f.tc ! //.-' : j.'jy j j/.j ! 3:sa
4 • ?:3o \ il.'f ' 3Wf <3c.3 ' v:.^7
5 i Afjv i //. f ' *+:3S £C.<, \ ^;vc,
: \ ••
RKOrdlnq STIOKI clcctranic Audit Utinq RtCinaei Si«ul>
A. sti?-:tv.i t«st
?.«ftrt^.c«
>:J?.IL Strip 0.-)i:«i
?•«: Ko. I.T14 2r .IT'.*) raadin^ r«adin9
; -^f 	 ^ — ^f-
, :o.o it.q wr
, is.c sa.3 s.s.(.
s //.o -jf.tf h\«r*\ o>'ic< loce-n'
2-rrc bet^ren 3-^/0 rl^rr a«rf .-to;(fet( /xr/tteat



	

\ Opacity
•*.,





General Syste* Coaaents i'.':ousff>ceepi^7 ciservacicns, cperacor sAili -eve is,
recommended corrections, ;jrprovenen cs ;
before oudjJc . 5.*% *'<& M£. rftuUv n^lec-tci m at
dtevtc*. r&uUnfS k- (^IdFftlS irtQ OfXLCL~C
-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 42
    Sufficient  data  should  be  taken  to  calculate  mean  errors,
confidence  intervals,  and  the  like.    Detailed  methods  for  these
calculations have been documented.2

9.3.2.4  Corrective Maintenance (Problems and Troubleshooting).  Even
in the  best of systems,  problems will eventually  develop.   Quality
control technique and quality assurance programs can help to minimize
problems,  but  they  cannot eliminate  them.   Table  9.5  lists  some
common and uncommon problems associated with transmissometer systems.
Faults  and  component  failures  are  generally   immediately  obvious.
Some problems can be so subtle that they will be detected only during
a period  of corrective  maintenance or  during  a  performance  audit.
But then  again,  some otherwise' obvious  problems  are  not recognized
for  long  periods  of  time  and  are  discovered  only  during  the
independent performance audit.

    Problems  that  are  site-specific  or   instrument-specific  may
develop at  an  installation.   The problem and  the corrective  action
taken should be described  in  the  logbook.   If  a recurring problem is
preventable, the method used to prevent it or the means of corrective
action  should  be  incorporated  into  the  preventive  maintenance
schedule.

    Troubleshooting  is  an  acquired  skill.     A  good  instrument
technician can piece together  a  number of clues and rapidly identify
a problem.   If the monitoring  system  is  new and  unfamiliar,  it may
take longer, but  the  time required should  decrease as experience is
gained.

    One of the most useful troubleshooting tools for transmissometers
is the  strip chart  recorder.   A  common tendency  is to eliminate the
strip chart  recorder with  the  rationalization that a computer  can be
programmed  to  identify  system  faults.     This   can   be  done,  but
computers  and  programming have  their own  faults.   In any case,  a
recorder provides a good backup to any data handling system, since it
is easier  to detect  trends from  the  analog output.   At  a glance,
noise levels and system performance can generally be evaluated.   The
strip chart  record  can indicate  a  number of the  problems  that have
been identified  in  Table 9.5.   Figure 9.10 illustrates  the type of
strip chart traces that might be observed.7

    The problems  that  can be  detected vary from environmental ones,
such  as  temperature  effects  and  vibration,   to  instrumentation
problems, such as sticking zero mirrors.

    Figure  9.10 presents  only a few examples of  clues that can help
the instrument  serviceman.   Microprocessors incorporated  in  modern

-------
                                                            Section  No.  3.0.9
                                                                    Date  6/1/86
                                                                        Page 43
             TABLE  9.5.   Opacity Monitoring System Problems
               Problem
           Corrective Action
Part 1.   Common Operator
Problems

Rezeroing and calibrating without
   cleaning windows.
Excessive zeroing  and  calibrating
   (attempts to adjust random noise).
Wrong stack-exit correlation value
   set in instrument.

Alarm goes off when system goes
    through span check.

Lack of correspondence between
   (1) meter opacity and optical
       density readings;
   (2) meter scales; or
   (3) meter, strip chart, and
       digital printout values.
Failure to clean windows when fault
   light indicates cleaning required.
   Failure to clean retroreflector
   window.
Improper use or no use of combiner
 .  equations for multiple monitor
   system.
Part 2.   Physical  Problems

Blower motor bearings  freeze-bearing
   noise.
Clean windows  first.

Set statistical  criteria  for making
   adjustments.
Correct calculation.   Flange-to-flange
   distance ofjten  mistakenly used rather
   than inside stack diameter.
Reprogram system.   A common programming
   error - although this  is sometimes done
   intentionally to check the alarm system.
Recalibrate system.  Most systems can be
   adjusted to produce consistent readings.
   It may take some work,  but a good
   operator will see that it is done.
Clean windows.
Correct calculations.
Replace bearings  -  develop better
   preventive maintenance program.
Excessive dirt buildup on  windows.
Cyclic drift in signal unrelated  to
   plant performance  - due to ambient
   temperature changes.
Monitor reads 100%  opacity for long
   period of time;  protective shutter
   in place.
Clean filters.
Insulate protective hood  or  install
   temperature  conditioning  system about
   monitor.
Reset shutter or  troubleshoot purge air
   supply (once activated and in place,
   some shutters  have  to  be  manually
   reset).
                                (continued)

-------
                                                             Section No.  3.0.9
                                                                    Date  6/1/86
                                                                        Page  44
                          TABLE 9.5.   (Continued)
               Problem
           Corrective Action.
Part 2.   Physical Problems
(Continued)
Excessive duct or stack  vibration
   causing misalignment  and
   electrical noise.
Improper alignment.
Part 3.   Electrical  Problems
Ground loops and noise.
Large voltage drops  when plant
   equipment is started;  spikes  in
   strip chart record.
Static electrical charges.

Lightning strikes burning electronic
   circuits.

System stuck in simulated zero or
   span calibration  mode.
Lamp out, erratic lamp output (a
   fault light should indicate this
   problem).
Mo output from instrument, no
   calibration cycle,  etc.
Improper instrument  responses -
   faulty calibrations,  improper or
   no outputs.
A definite problem.   One  solution  is  to
   mount transceiver  and  retroreflector
   assemblies independently  from this  stack
   or duct.   Make connections  between  duct
   and monitor using  a  flexible bellows.
   Another is to  move the monitor  to  a less
   vibration-prone location.
Realign, check, and  tighten  system.


Trace and rewire.
Install transient suppressor,  dedicated
   power transformer  for  monitoring system.

Connect transmissometer case to dedicated
   earth ground.
Add phenolic gaskets  between metal stack
   and transmissometer.   Add surge
   arrestors to junction  at  junction  box.
Check solenoids and motors  in  transceiver
    clean or replace.
Check modulator and motor; adjust or  re-
   place motor if necessary.   Check lamp
   and replace; when  replacing, keep  glass
   surfaces of lamp clean; avoid finger-
   prints and clean with  lens  solution
   before turning on.
Check fuses (hope that  it is this  simple).

Check electronics. Check to see that  cards
   and components are secure.  Use
   troubleshooting guide  supplied  by  vendor
   to check electronic  test  points.
   Replace appropriate  components  or
   replace cards.

-------
                                                       Section No.  3.0.9
                                                             Date  6/1/86
                                                                 Page  45
instrument  systems are  today  being programmed  to directly  alert  the
operator  about such problems.   However,  a computerized  system can  be
programmed  to identify  only  a  set  number  of  malfunctions.     If
problems  occur  that  are  not  identified  by  such a  system,  a  strip
chart recorder can be a valuable tool in uncovering  them.
                           Suspect
                                               Excessive drift
                                               Window fouling
                   Straight line trace
                   sustained for several
                   hours
Cyclic pattern/
temperature sensitivity l?l
                   Spikes-
                   electrical instability
                Source: Larkin, R.. 1977. Ref.7.
Drift, gradually
increasing or
decreasing
                                               Bad zero check
                   Figure 9.10.  Possible strip chart
                       traces indicating problems.
9.4  References

1.   Code  of  Federal  Regulations.    Title  40   Protection  of
     Environment.  Part 60 Appendix B.  40 CFR 60  App.  B, 1985.
                                     the
2.   Jahnke,  J.A.  and Aldina,  G.J.   Continuous  Air  Pollution  Source
     Monitoring Systems.   EPA 625/6-79-005, June  1979.

3.   Traceability  Protocol  for  Establishing  True  Concentrations  of
     Gases  Used  for  Calibration   and   Audits  of  Continuous   Source
     Emission Monitors  (Protocol Number  1).   June  1978,  Section  3.0.4
     of  the  Quality Assurance  Handbook  for  Air  Pollution Measurement
     Systems,  Volume  III,  Stationary  Source  Specific   Methods.
     EPA-600/4-77-027b.   August  1977.   U.S.  Environmental Protection
     Agency,  Office  of  Research  and Development  Publications, 26  West
     St.  Clair Street, Cincinnati,  OH  45268.

-------
                                                Section No. 3.0.9
                                                      Date 6/1/86
                                                          Page  46
Code  of  Federal  Regulations.    Title  40  Protection  of  the
Environment.  Part 60.13, Monitoring Requirements.  40 CFR 60.13,
1985.

Standards of Performance for New  Stationary Sources:   40 CFR 60,
Appendix F - Quality  Assurance Procedures,  Procedure  1 - Quality
Assurance Requirements for Gaseous Continuous Emission Monitoring
Systems Used for Compliance Determination.

Purcell,  R.Y.   and  Rosenquest,  J.M.    Field Performance  Audit
Procedures  for  Opacity  Monitors.   EPA CEM  report   series  No.
5-271-7/82  (see also, Performance  Audit Procedures  for Opacity
Monitors.  EPA  340/1-83-010), 1982.

Larkin,  R.,  Jaye,  F.,  and Steiner,  J.    Resource   Manual  for
Implementing the NSPS  Continuous  Monitoring Regulations.  Manual
4 -  Source Operating  and  Maintenance Procedures  for Continuous
Monitoring Systems.   EPA 340/l-78-005d,  1978.

Additional References

Entropy  Environmentalists,  Inc.   Assessment  of the  Adequacy of
the Appendix F  Quality Assurance  Procedures for Maintaining CEMs
Data Accuracy:   Status Report #1.   EPA 600/4-83-047,  1983.

Entropy Environmentalists, Inc.   Guidelines for the Observations
of  Performance Specification  Tests   of  Continuous  Emission
Monitors.  EPA  340/1-83-009,  1983.

Jahnke, J.A.   APTI  Course SI:   476A   Transmissometer Systems -
Operation and Maintenance. An Advanced Course.  EPA 450/2-84-004,
September 1984.

Logan,  J.T.  and Rollins,  R.   Quality  Assurance  for Compliance
Continuous Emission Monitoring  Systems:  Evaluation of Span Drift
for  Gas  CEMs,   Proceedings  - APCA/ASQC   Specialty  Conference:
Quality Assurance in  Air  Pollution Measurements.   October 10-14,
1984, Boulder,  CO.

Osborne,  M.C.   and  Midgett,  M.R.    Survey  of  Continuous  Source
Emission Monitors:   Survey No.  1  NOx and SO2.  EPA 600/4-77-022,
1977.

Peeler, J.W.   A Compilation of SO2 and NOy  Continuous Emission
Monitor Reliability Information.  EPA 340/1-83-012,  1983.

-------
                                                    Section No. 3.0.9
                                                          Date 6/1/86
                                                              Page 47
7.  Peeler, J.W.  Performance Audit Procedures for S02r NOx/ CO2, and
    02 Continuous  Emission  Monitoring  Systems.   EPA 340/1-85-015,
    1983.

8.  Reynolds,  W.E.    Development  and  Evaluation  of  SO2  CEM  QA
    Procedures.    Proceedings  -  APCA/ASQC  Specialty  Conference:
    Quality Assurance in Air  Pollution  Measurements.   October 10-14,
    1984, Boulder,  CO.

9.  Wayne, A.   Opacity Monitoring  Quality  Assurance  Programs in U.S.
    EPA  Region VIII.    Proceedings-APCA/ASQC Specialty  Conference:
    Quality Assurance in  Air Pollution Measurement.   October 10-14,
    1984,  Boulder, CO.

-------
                                           Section No. 3.0.10
                                           Date November 26, 1985
                                           Page 1

10.0  GUIDELINE FOR DEVELOPING QUALITY CONTROL PROCEDURES
      FOR GASEOUS CONTINUOUS EMISSION MONITORING SYSTEMS

10.1  Introduction

    This guideline describes the minimum  content  for  a quality
control  plan  to  satisfy  the  requirements  of  Section  3  of
Appendix F, Procedure  1  to  40  CFR  Part 60.  Source owners or
operators  may wish to add other items to assure  the  generation
and  reporting  of  valid  data  from  their  continuous emission
monitoring systems (GEMS's).

    Appendix F, Procedure 1 requires written  procedures for each
of the following activities:

    1.   Calibration of the CEMS.

    2.   Calibration  drift  determination  and adjustment of the
         CEMS.

    3.   Preventive maintenance of the CEMS  (including maintain-
         ing a spare parts inventory).

    4.   Data recording, calculations,  and  reporting  for emis-
         sions and QA data.

    5.   Accuracy  audit  procedures   including   sampling   and
         analysis methods.

    6.   Program  of  corrective  action  for  the malfunctioning
         CEMS.

    Figure  1  is  a  flow  chart  showing  the  requirements  in
Appendix F, Procedure 1 for quality assurance  and  in Part 60.13
for monitoring requirements.  This flow chart is included to show
how these requirements for CEMS's interact.

10.2  Calibration of the CEMS

    Calibration refers  to  the  adjustment  of the CEMS response
relative to specified  standards such as gas cells or calibration
gases,   or   relative   to  independent  effluent  measurements.
Appendix  F,  Procedure  1  requires  that  sources  have written
procedures for CEMS calibration.  Sources  may  develop their own
written procedures; alternatively,  they  may  specify applicable
sections of the instrument manual as their written procedures.

    There  are  no currently promulgated regulations that require
either specific calibration  frequencies or specific criteria for
initiating calibration procedures.  Sources may  therefore choose
their  own  frequency  or  criteria   for  calibration  based  on
operating experience or manufacturer's recommendations.

-------
                                                   Section No.  3.0.10
                                                   Date  November  26,  1985
                                                   Page  2
                                        START
    DAILY DRIFT CHECK
               PERFORM DAILY
                CAL. DRIFT
               DETERMINATION
       5TH
 N / CONS.
     DAY CD >
     (2MSPEC)
                             PASS
DECLARE CEMS OUT-
   OF-CONTROL
 BEGINNING AT END
  OF 5TH 24-HR.
     PERIOD
 DECLARE CEMS
 OUT-OF-CONTROL
BEGINNING AT END
  OF PREVIOUS
 SUCCESSFUL CD
    CHECK
     PERFORM
 CORRECTIVE ACTION
 REPEAT CD CHECK;
  WAIT UNTIL TIME
  FOR NEXT DAILY
    CD CHECK
    PERFORM
CORRECTIVE ACTION
  REPEAT CD CHECK;
  WAIT UNTIL TIME
  FOR NEXT DAILY
     CD CHECK
                                             COLLECT DATA
                                               FOR DAR
    DECLARE CEMS BACK IN CONTROL
     BEGINNING AT COMPLETION OF
      LAST SUCCESSFUL CD CHECK
                    INVALIDATE DATA
                    FROM BEGINNING TO
                    END OF O.O.C. PERIOD.
                    DESCRIBE CORRECTIVE
                   .ACTION FOR DAR
         Figure  1.    Flow  Chart  for Required  QC  Procedures.

-------
                                                                    Section No.   3.0.10
                                                                    Date  November  26,  1985
                                                                    Page  3
                                                                                         WAIT UNTIL TIME
                                                                                         FOR NEXT DAILY
                                                                                         CAL DRIFT CHECK
                                                                           TIME
                                                                       FOR REQUIRED
                                                                        PREVENTIVE
                                                                       MAINTENANCE
                                                                              TIME
                                                                           FOR OTRLY
                                                                            ACCY. CK.
                                                                               1
                                                                     CONDUCT REQUIRED
                                                                        PREVENTIVE
                                                                        MAINTENANCE
QUARTERLY
ACCURACY
AUDIT
RECORD ACCURACY RESULTS FOR DAR. IF CEMS
FAILED INITIAL ACCURACY TEST THEN:
 (A) DECLARE CEMS O.O.C. AND INVALIDATE
   CEMS DATA FROM END OF SAMPLING FOR
   INITIAL TEST UNTIL END OF SAMPLING
   FOR SUCCESSFUL ACCURACY TEST
 (B) REPORT DATA FROM INITIAL AND
   SUCCESSFUL ACCURACY TESTS
 (C) DESCRIBE CORRECTIVE ACTION
    CHANGE QAP
       OR
    REPAIR CEMS
       OR
    REPLACE CEMS
                    2ND
                CONSECUTIVE
                 QTR. WITH
                UNACCEPTABLE
                ACCY. RESULTS
              CODE:
                    £    - "EQUAL TO OR LESS THAN"
                    >    . "GREATER THAN"
                    CD   - CALIBRATION DRIFT
                    CONS - CONSECUTIVE
                    DAR   - DATA ASSESSMENT REPORT
                                         N     - NO
                                         O.O.C.  - OUT-OF-CONTROL
                                         QCP   - QUALITY CONTROL PROCEDURES
                                         SPEC  - DRIFT LIMITS IN PERFORMANCE
                                                  SPECIFICATIONS 2 OR 3
                                         Y     . YES
                                           Figure  1.    (continued)

-------
                                           Section No. 3.0.10
                                           Date November 26, 1985
                                           Page 4

    For calibrations based on external gas cells, sufficient time
should be allowed for the cell and/or analyzer  cabinet  to reach
normal operating temperature; accordingly, it is recommended that
procedures  be  incorporated  into  the  QC  program  that ensure
sufficient time for the monitor  response  to stabilize before it
is compared to the cell's named value.   Some  in-situ  analyzers
partially  or totally disable temperature compensation  circuitry
during cell-type calibrations.  In these cases, it is recommended
that  additional  procedures addressing the calibration  of  this
circuitry be incorporated into the QC program.

    For  analyzers  calibrated  using  calibration  gases  as the
reference, the written procedures  should  specify  (1)  at  what
point  in  the sampling system the calibration gases  are  to  be
introduced and (2) either the specific  gas  flow rate to be used
or how the flow rate  is determined.  Although current continuous
emission monitoring (CEM) regulations do not require establishing
the traceability of calibration  gases to higher standards, it is
strongly recommended that procedures be established and  included
within the QC program for verifying the concentrations  of  cali-
bration gases.  One acceptable  procedure  is  EPA's traceability
protocol 1 (Reference 1).

    In cases where a portable CEMS is to be used as the reference
for  adjusting  the installed  CEMS,  written  procedures  should
specify  calibration  and  operating  procedures for the portable
CEMS, including the portable CEMS sampling location.

    The written calibration procedures for the installed CEMS may
be incorporated into one or more of the following  sections  of a
QC program:

    1.   A Stand-Alone "Calibration" Section  of  the QC Program.
         In  this  case, the  frequency  of  calibration  or  the
         criteria for initiating calibration activities should be
         clearly specified.

    2.   Preventive Maintenance.  Within the  section delineating
         the preventive maintenance  procedures,  calibration may
         be  specified  as a routine maintenance activity  to  be
         performed  at  regular,  specified  intervals.  Alterna-
         tively,  calibration  may  be  specified on an as-needed
         basis with stated  criteria  for  the  implementation of
         calibration activities.

    3.   Corrective   Action.    Calibration  procedures  may  be
         included   within  the  section  delineating  corrective
         action  activities  to be performed at the discretion of
         CEMS repair personnel  in  response to an out-of-control
         CEMS.

Regardless  of  how the calibration procedures  are  incorporated
into the QC program; it is recommended  that  the  individual  or

-------
                                           Section No. 3.0.10
                                           Date November 26,  1985
                                           Page 5

group responsible for  GEMS  calibration be identified within the
written QC plan.

10.3  Calibration Drift and Adjustment of the CEMS

    Calibration drift (CD) refers to the difference  between  the
CEMS output reading and a  reference  value  after  a  period  of
operation  during  which  no  unscheduled maintenance, repair, or
adjustment took place.  Daily zero (or low  level) and span drift
checks are required by 40 CFR 60.13; these checks  are to be used
to fulfill the calibration drift check requirement of Appendix F,
Procedure 1.  Appendix F, Procedure 1 requires written procedures
that  specify  how the zero (or low level) and  span  calibration
drift determinations are to be performed.  These  procedures must
be consistent with the monitor  vendor's  prescribed  method  for
checking CD.

    Table   10.1  presents  CD  criteria  and  the  corresponding
required source responses. Sources may choose to  establish  more
stringent criteria for adjustment of CEMS for zero (or low level)
and/or span calibration  drift.   It  is  recommended that the CD
criteria selected for adjustment of the CEMS be incorporated into
the  written   instructions   for  the  calibration  drift  check
procedures, so that the need for adjustment based  on calibration
drift may be determined immediately.

    Corrections  for  excessive   drift   may   consist   of  any
adjustments  or  activities that the operator or technician deems
necessary to correct for the observed  drift.   These  activities
typically  consist  of  routine   checks   and   adjustments   of
calibration gas flow rates and pressures,  verification of proper
sample cell temperatures, verification  of  the status of monitor
specific auxiliary monitoring  parameters, and adjustment of zero
and/or  span  potentiometers.   Written  procedures   should   be
available for  performing  these  routine  activities  and should
include  criteria  for  determining  that  adjustments  have been
successful.

10.4  Preventive Maintenance of the CEMS

    Preventive maintenance is comprised of activities designed to
detect and prevent the development of monitoring problems.  These
activities typically include both  routine maintenance procedures
and maintenance, repairs,  or  adjustments  performed  on  an as-
needed basis.  An example  of  as-needed  preventive  maintenance
would  be  the  repairing  of  the  protective  covering   of  an
extractive  sample  line  following  damage  resulting  from   an
accident during the construction activities.   If the sample line
itself  were   not   damaged,  the  repair  would  be  considered
preventive  maintenance  and  would  not  consititute  corrective
action for a malfunctioning CEMS.  The importance of this type of

-------
                                           Section No. 3.0.10
                                           Date November 26, 1985
                                           Page 6
           TABLE 10.1.  CEMS CALIBRATION DRIFT CRITERIA
Parameter
  Criterion*
     Action Required
Zero (or low)
level cali-
bration drift
CD > 2 x (Spec)**
                 CD > 2 x (Spec) for
                 5 consecutive 24-hour
                 periods
                 CD > 4 x (Spec)
Adjust CEMS for calibra-
tion drift
                        CEMS out-of-control period
                        begins at end of 5th day
                        the CD exceeds 2 x (Spec);
                        perform corrective action
                        and repeat CD check

                        CEMS out-of-control period
                        begins at the time corres-
                        ponding to the completion
                        of the last acceptable CD
                        check preceding the CD
                        check which exceeds
                        4 x (Spec); perform
                        corrective action and
                        repeat the CD check
Span cali-
bration drift
CD > 2 x (Spec)**
                 CD > 2 x (Spec) for
                 5 consecutive 24-hour
                 periods
                 CD > 4 x (Spec)
Adjust CEMS for calibration
drift

CEMS out-of-control period
begins at end of 5th day
the CD exceeds 2 x (Spec);
perform corrective action
and repeat CD check

CEMS out-of-control period
begins at the time corres-
ponding to the completion of
the last successful CD check
preceding the CD check that
exceeds 4 x (Spec); perform
corrective action and repeat
the CD check
 *Spec refers to the applicable performance specification in
  Appendix B.

**This is the minimum criterion for adjustment of the CEMS.  More
  stringent criteria, which may be preferred by many sources, are
  also acceptable.

-------
                                           Section No. 3.0.10
                                           Date November 26, 1985
                                           Page 7

maintenance is recognized;  however,  it is neither practical nor
necessary  to  develop  written  procedures   for   such   needed
activities.

    Written procedures must be  available for routine maintenance
activities.  These procedures should  specify what procedures are
to  be  conducted  and  the  frequency  with  which  the  various
activities  are  to  be performed.  The QC program should specify
the individual  or  office  responsible  for  ensuring  that  the
preventive   maintenance   procedures   are   conducted   at  the
appropriate  frequencies  and  the individual or group  who  will
perform the actual routine maintenance procedures.

    The applicable regulations  do  not specify the minimum level
of routine preventive maintenance.  It is suggested  that,  at  a
minimum,  the  initial procedures should incorporate the vendor's
recommendations  regarding  preventive maintenance activities and
frequencies.  These procedures  may  later be adjusted to reflect
actual operating experience with individual CEMS installations.

    A list of spare parts for the CEMS  must  be  included in the
written QC  plan.  At a minimum, those spare parts recommended by
the monitor  vendor  should  be available.  The QC program should
specify  the  individual  or  office  who   is   responsible  for
maintaining the listed spare parts inventory.

10.5  Data Records, Calculations, and Reporting for the CEMS

    The  QA/QC  program must address recordkeeping, calculations,
and  reporting of emissions  and  quality  assurance  data.   The
requirements  for  these activities are contained in the subparts
of  40  CFR  60  that  specify the use of CEM.  A Data Assessment
Report (DAR) must be provided with emissions reports  required by
the applicable subpart of 40 CFR 60.  The DAR must contain, at  a
minimum:

    1.  The name and address of the source owner or operator.

    2.  Identification and location of each monitor in the CEMS.

    3.  The manufacturer and model number of  each monitor in the
        CEMS.

    4.  Quarterly  accuracy   results,   including   dates,  CEMS
        responses,  and  either   reference   method  results  or
        certified  gas  values;  if  either  a  RATA or a RAA was
        performed,  the  results  from  the EPA performance audit
        sample analysis must also be included.

    5.  A summary  of  corrective  actions taken when the monitor
        was determined to be out-of-control.

-------
                                           Section No. 3.0.10
                                           Date November 26, 1985
                                           Page 8

    For emissions  data,  a  list  or  diagram should be provided
indicating   the  offices  or  individuals  responsible  for  (1)
retrieving the  data  from  the  CEMS,  (2) calculating emissions
rates from the CEMS  data,  (3)  compiling emissions reports, and
(4) reviewing and/or approving  emissions  reports.  Formulas and
example   calculations  should  be  provided  for  emission  rate
calculations.   Similar   information   should  be  provided  for
emissions  data  from  alternative monitoring methods that may be
necessary during CEMS out-of-control periods.

    A  list  or  diagram should also be provided  indicating  the
offices  or  individuals responsible for (1)  collecting  quality
assurance (QA) data, (2) performing  applicable  calculations  of
QA/QC  results,  (3)  recording  the QA/QC results in appropriate
logs (as applicable),  (4)  preparing  the DAR, and (5) approving
and/or  reviewing  the  DAR.   Formulas and example  calculations
should be provided for all required QA data calculations.

10.6  Accuracy Audit Procedures Including Sampling and Analysis
      Methods

    Appendix F, Procedure 1 requires that each CEMS be audited at
least once each calendar quarter.   Three  audit  techniques  are
acceptable:

    1.  Relative accuracy test audits (RATA's);

    2.  Cylinder Gas Audits (CGA's); and

    3.  Relative accuracy audits (RAA's).

In addition, other alternative  audit  procedures  may be used as
approved by the Administrator.

    If the  CEMS  does not demonstrate acceptable accuracy during
the quarterly audit, then corrective actions  must  be initiated,
and the  CEMS  must be declared out-of-control from the time cor-
responding to the completion of the sampling for the unsuccessful
audit  until  the  completion of the sampling  for  a  successful
follow-up audit.  If the CEMS  demonstrates unacceptable accuracy
for  two  consecutive quarters,  then  the  QA  program  must  be
revised, or the CEMS must be modified or replaced.

    Table  10.2  presents  the  specific  requirements   and  the
corresponding CEMS performance  criteria  for  each  of the three
acceptable audit techniques.

    The  QC  program  must  include written sampling and analysis
procedures  to  be  used during the required  quarterly  accuracy
audits.  At a minimum, these procedures must describe the methods
to be used to  conduct a RATA.  Applicable sections of Appendix A
(Reference Methods)  and  Appendix B (Performance Specifications)

-------
                                           Section No.  3.0.10
                                           Date November 26, 1985
                                           Page 9
            TABLE 10.2.  REQUIREMENTS AND CRITERIA FOR
             APPENDIX F, PROCEDURE 1 AUDIT TECHNIQUES
Technique
  Requirements
  Performance Criteria
RATA
RAA
CGA
Conduct as per
applicable performance
specification (PS) in
Appendix B (e.g., PS 2
for S09 and NO )
      £»       Ji

Analyze appropriate
performance audit
samples from EPA
RA must not exceed 20% or
10% of applicable standard,
whichever is greater
                                      For SO,
        standards from
Conduct as per appli-
cable PS in Appendix
B except only 3 runs
are required

Use relative difference
between the mean
reference method values
and the mean of the CEMS
responses to assess the
accuracy of the CEMS data

Challenge both pollu-
tant and diluent chan-     H
nels (if applicable)
of CEMS three times at
the two points specified
in Procedure 1

Use gases that have been
certified by comparison
to NBS SRM's or NBS/EPA
approved gas manufacturer's
CRM's

Operate analyzer in normal
sampling mode
           Use average difference between
           actual gas value and concentra-
           tion indicated by CEMS to
           access accuracy
0.20 to 0.30 Ib/lO" Btu,
RA must not exceed 15% of
the standard

For S09 standards below
0.20 IB/10  Btu, RA must
not exceed 20% of the
standard

Inaccuracy must not exceed
+ 15% or 7.5% of the appli-
cable standard, whichever
is greater
Inaccuracy must not exceed
+ 15%

-------
                                           Section No. 3.0.10
                                           Date November 26, 1985
                                           Page 10

may  be  cited  where possible to describe audit procedures.  The
written procedures should specify individuals or  groups  respon-
sible  for audit program oversight, sampling, analysis, and accu-
acy assessment calculations.  If the source  chooses  to  conduct
RAA's and/or CGA's during  quarters when RATA's are not required,
the QC plan should include  written  procedures  for  these audit
techniques.  Again, applicable  sections  of  Appendix  A, Appen-
dix B, and/or instrument  operation  manuals  may  be cited where
possible.

    Sources may choose to have an outside contractor perform some
or all of the accuracy audit activities.  Since contractor selec-
tion may be subject to competitive  bidding,  the QC program need
not specify  a  particular  contractor.   However,  the  specific
activities for which the contractor will be responsible should be
listed.

10.7  Program of Corrective Action for the Malfunctioning CEMS

    Appendix F, Procedure 1 specifies that corrective action must
be   performed   when  a  CEMS  is  out-of-control.   Appropriate
corrective   action  will  depend  on  the  nature  of  the  CEMS
malfunction.  At a minimum, written procedures must be available,
to  be  applied as necessary, for instrument start-up and trouble
shooting.   Appropriate  sections of instrument operation  and/or
repair manuals  may  be  referenced - to fulfill this requirement.
Where  possible,  it  is  recommended  that   additional  quality
assessment  procedures  be provided to verify proper operation of
the CEMS following repair or adjustment.

    A  list  should  be provided  to  indicate  what  alternative
methods  are  to  be used for monitoring  emissions  during  CEMS
out-of-control  periods  in  order  to  fulfill the minimum  data
availability  requirements  of  the applicable subpart.   Written
procedures should be available for operation of these alternative
methods.

    A list or chart should be provided to indicate the offices or
individuals (1) to be contacted when a CEMS out-of-control period
occurs, (2) to approve the corrective action (if applicable), and
(3 j to be responsible for determining when alternative monitoring
procedures  are to be employed.  Criteria should be provided  for
determining when the CEMS is out-of-control.  As a minimum, these
must include  the  Appendix F, Procedure I criteria for excessive
drift and excessive inaccuracy.

10.8  References

   1. Traceability  Protocol for Establishing True Concentrations
      of  Gases Used for Calibration  and  Audits  of  Continuous
      Source  Emission Monitors (Protocol Number 1).  June  1978,
      Section  3.0.4  of the Quality Assurance Handbook  for  Air

-------
                                     Section No. 3.0.10
                                     Date November 26, 1985
                                     Page 11

Pollution   Measurement  Systems,  Volume,  ill,  Stationary
Source Specific Methods.  EPA-600/4-77-027b.   August 1977.
U. S. Environmental Protection Agency,  Office  of Research
and  Development  Publications,  26  West St. Clair Street,
Cincinnati, Ohio   45268.

-------
                                                  Section No.  3.14
                                                  Date July 1,  1986
                                                  Page 1
                         Section 3.14
         METHOD 7A - DETERMINATION OF NITROGEN OXIDE
              EMISSIONS FROM STATIONARY SOURCES

         (Grab Sampling - Ion Chromatographic Method)
                           OUTLINE


                                                      Number of
  Section                            Documentation      pages

SUMMARY                                   3.14             2

METHOD HIGHLIGHTS                         3.14             8

METHOD DESCRIPTION

   1.  PROCUREMENT OF APPARATUS
       AND SUPPLIES                       3.14.1          10

   2.  CALIBRATION OF APPARATUS           3.14.2          14

   3.  PRESAMPLING OPERATIONS             3.14.3           6

   4.  ON-SITE MEASUREMENTS               3.14.4           7

   5.  POSTSAMPLING OPERATIONS            3.14.5          11

   6.  CALCULATIONS                       3.14.6           6

   7.  MAINTENANCE                        3.14.7           2

   8.  AUDITING PROCEDURES                3.14.8           6

   9.  RECOMMENDED STANDARDS FOR
       ESTABLISHING TRACEABILITY          3.14.9           1

  10.  REFERENCE METHOD                   3.14.10          3

  11.  REFERENCES                         3.14.11          2

  12.  DATA FORMS                         3.14.12         12

-------
                                                    Section No. 3.14
                                                    Date July 1, 1986
                                                    Page 2
                               SUMMARY
    A gas sample is extracted from the  sampling  point in the stack.
The  sample  is collected in an evacuated 2-liter round bottom  boro-
silicate  flask containing 25 ml  of  dilute  sulfuric  acid-hydrogen
peroxide  absorbing  reagent.  The nitrogen oxides, NO and NO^, react
with  the absorbing reagent to form nitrate ion which is analyzed  by
ion chromatography (1C).  The  method  does  not  respond  to nitrous
oxide, N20.

    The reactions  that  describe  absorption of the NO  are distinct
for NO and N02._ The common feature of the reactions isxthe formation
of nitrate, N03~, as nitric acid, HN03.

    The  absorption  of  NO involves an oxidation-reduction  reaction
where the  oxidizing  agent is the acidic hydrogen peroxide solution.
The two half reactions are:
              3H2°2 + 6H+ + 6e~ = 6H2°
and
        2NO + 4H-0 = 2NO ~ + 9H+ + 6e~;
                £       O

and the overall reaction is:

                         • —     +
        O KT^ _i_ O U f\  _ O VT/^   _i_ O tl  _i_ O U i
        zNO -f on—O-t = ^JNO—  + 2.n  + 2H~<
    The absorption of N02 presumably involves the reaction with water
to  form  nitric  acid and NO.  NO, reacts with water to form  nitric
acid and nitrous acid, HN02:

              2N02 + H20 = HN03 + HN02.

    The nitrous acid is unstable and decomposes:

              3HNO2 - 2NO + HN03 + H20.

    The observed reaction is the sum of the two reactions above:

              3NO2 + H2O = 2HN03 + NO.

    Absorption of N02 proceeds  faster  than absorption of NO because
N0«  is  more  soluble in solution, where reaction occurs.   In  this
respect, it should be noted  that  absorption of NO is quickened as a
consequence of reaction with oxygen also present within the flask:

              2ND + 02 = 2N02.

   If the gas being sampled  contains  insufficient  oxygen  for  the
conversion of NO to N02, then  oxygen  should  be introduced into the
flask by  one  of  three  methods: (1) before evacuating the sampling

-------
                                                    Section No. 3.14
                                                    Date July 1, 1986
                                                    Page 3

flask,  flush it with pure cylinder oxygen,   and  then  evacuate  the
flask to 75 mm (3.0  in.) Hg absolute pressure or less; or (2) inject
oxygen into the flask after sampling; or (3) terminate  sampling with
a minimum of 50 mm (2.0 in.) Hg vacuum remaining in the flask, record
this final pressure, and then open  the flask to the atmosphere until
the flask pressure is almost equal to atmospheric pressure.

    Method 7A is applicable to  the  measurement  of  nitrogen oxides
emitted from stationary sources.  It may be used as an alternative to
Method 7 (as defined in 40 CFR Part 60.8(b)) to determine  compliance
if  the  stack  concentration  is within the analytical  range.   The
analytical range of the method is from 125 to 1250  mg NO , expressed
as  NO?'  per  drv  standard  cubic meter (65 to  655  ppm).   Higher
concentrations  may  be  analyzed by diluting the sample.  The  lower
detection limit is approximately  19  mg/m   (10  ppm),  but may vary
among instruments.

    The method description  which follows is based on the method that
was promulgated on'December 8, 1983.

    Section  3.14.10  contains  a  copy of Method 7A,  and blank data
forms are provided in Section  3.14.12  for  the  convenience  of the
Handbook user.

    Note:   Because  of  similarities  between Method 7A and Method 7
sampling equipment and procedures, in most cases only the differences
in Method 7A are presented in  detail  in  this section (3.14).  How-
ever, all tasks are shown in the activity  matrices  and  data sheets
needed to perform  Method 7A are included, whether or not differences
occur in the written descriptions.   Other  Method  7A procedures are
referenced  to  the  corresponding description in Section 3.6, Method
7.  This is done for both time savings to the reader and cost savings
to the Government.

-------
                                                    Section No. 3.14
                                                    Date July 1, 1986
                                                    Page 4
                          METHOD HIGHLIGHTS
    Section  3.14  (Method  7A)  describes the required procedure for
sampling and analyzing of nitrogen oxides emissions  from  stationary
sources.   For the method, a grab sample is extracted from a point in
the stack, and collected in a previously evacuated flask containing a
sulfuric   acid-hydrogen   peroxide  absorbing  solution.   With  the
exception  of  nitrous  oxide, the nitrogen oxides  are  oxidized  to
nitrate which is analyzed by ion chromatography  (1C).   Results  are
expressed   as   concentrations   of  nitrogen  dioxide  (N02).   The
applicable  regulation  should  be  consulted  to  determine  whether
additional  measurements, such as velocity or CU determinations,  are
required.

    The  absorbing reagent for EPA Method 7A has a sulfuric acid con-
centration  one-tenth  that  of EPA Method 7.  In all other respects,
the sampling train and sampling procedures of EPA Method 7A are iden-
tical to those of  EPA  Method  7.   Sample preparation involves only
dilution  to  reach  a  measurable concentration range  for  the  ion
chromatograph.

    Ion chromatography  is  a  relatively  recent analytical develop-
ment.   The  reader  is  referred,2to  the  literature  for  detailed
descriptions  of  the  subject.      Small, et  al.,   developed  the
technique  using the principles of ion  exchange  chromatography  and
conductimetric  detection.   Previous  attempts  to  use this type of
detection were unsuccessful because of the presence of the background
electrolyte used for  elution  of  the ionic species.  Small, et al.,
used a novel combination of resins  to  separate the ions of interest
and neutralize the eluent from the background.

    The  aqueous sample is introduced into a fixed-volume sample loop
by using a plastic  syringe.   Once  injected,  the sample is carried
through  a  separation  column  at different rates according to their
relative affinities for the resin and the  eluent  and  are therefore
separated into discrete bands.   The  separated  ions are then passed
through a post-separation suppressor device, a source of hydrogen ion
(H ), which converts the eluent ions into a less conducting weak acid
while converting the analyte  ions  into  a  highly  conducting form.
This permits  the  use  of  a  conductivity  cell as a very sensitive
detector of all ionic species.

    Gjerde, et al.,   described a modified ion chromatographic method
that eliminates  the  need  for  a  suppressor  device.   Anions  are
separated  on  a column containing an anion-exchange resin with a low
exchange capacity.   Because  of  the  low  capacity,  a  very dilute
solution of an aromatic organic acid salt may be used as  the eluent.
The conductance of the eluent is sufficiently low that no suppression
is needed.

    For Method  7A,  either  suppressed  or  non-suppressed 1C may be
used.    The   basic  ion  chromatograph  will  have  the   following
components:

-------
                                                    Section No.  3.14
                                                    Date July 1,  1986
                                                    Page 5

    (a)  sample injection device,

    (b)  anion separation column,

    (c)  anion suppressor column,  either packed bed or fiber
         type (not required for non-suppressed 1C),

    (d)  conductivity detector, and

    (e)  recorder.

    Two  critical aspects of Method 7A are (a) the measurement of the
gaseous  sample volume, and (b) the preparation  of  the  calibration
standards for the ion chromatograph.   Analysts are advised to observe
specified  procedures  carefully  at  these  points  of  the  method.
Analysts performing the  method  should be well trained in the use of
the ion chromatograph.

    Collaborative testing of EPA Method  7A  has  not been performed.
However, from a technical standpoint, it can  be  expected  that  EPA
Method 7A will  exhibit  accuracy  and  precision  as good as, if not
better than, EPA Method 7.

    The four  blank  data  forms  at  the  end of this section may be
removed from the Handbook and used in the pretest, test, and posttest,
operations.  Each form has a subtitle (e.g.,   Method  7A, Figure 3.1)
to assist the  user in finding a similar completed form in the method
description (Section 3.14.3).   On the blank and filled-in forms, the
items/parameters that can  cause  the  most  significant  errors  are
designated with an asterisk.

1.  Procurement of Apparatus and Supplies
    Section  3.14.1   (Procurement  of  Apparatus and Supplies)  gives
specifications,  criteria,  and  design  features  for  the  required
equipment and materials.  The sampling apparatus  for  Method  7A has
the same design features as that of Method 7.  Section 3.14.1 can  be
used as a guide for procurement  and  initial checks of equipment and
supplies.  The activity matrix (Table 1.1) at the  end of the section
is a summary of the details given  in  the  text and can be used as a
quick reference.

2.  Pretest Preparations
    Section 3.14.2  (Calibration of Apparatus) addresses the required
calibration  procedures and considerations for the Method 7A sampling
equipment (same as Method 7) and analytical equipment (the ion chrom-
atograph).  Required accuracies for each component are also included.
A pretest  sampling  checklist  (Figure  3.1  in Section 3.14.3) or a
similar form should be used  to  summarize  the calibration and other
pertinent pretest data.  The volume of each collection  flask must be
determined with  stopcock  in  place.   This  volume  measurement  is
required only on the initial calibration,  provided  the  stopcock is
not changed.   The  calibration section may be removed along with the

-------
                                                    Section No. 3.14
                                                    Date July 1, 1986
                                                    Page 6

corresponding sections  from  the  other  methods  and  made  into  a
separate quality assurance  reference  manual  for  use  by personnel
involved in calibration activities.

    Section 3.14.3 (Presampling Operations) provides  the tester with
a guide  for  equipment  and supplies preparation for the field test.
With the  exception of the preparation of certain reagents, these are
the same as for Method  7.   A  pretest preparation form (Figure 3.2,
Section  3.14.3)  can  be used as an equipment checkout  and  packing
list.  The flasks may be charged with the absorbing  reagent  in  the
base laboratory.   The method of packing and the use of the described
packing containers should  help protect the equipment, but neither is
required by Method 7A.

    Activity matrices for the calibration  of  equipment and the pre-
sampling operations (Tables 2.2 and 3.1) summarize the activities.

3.  On-Site Measurements
    Section  3.14.4   (On-Site  Measurements)  contains  step-by-step
procedures  for  sample  collection  and for sample recovery.  Sample
collections  are the same as for Method  7;  sample  recovery  proce-
dures differ slightly  from  Method  7 in that the sample pH does not
have  to be checked and adjusted.  The on-site checklist (Figure 4.3,
Section 3.14.4) provides the tester  with  a quick method of checking
the  on-site  requirements.   When high negative stack pressures  are
present, extra  care  should be taken to purge the leak-tested sample
system and to be sure  the  flask  is  £  75  mm  (3 in.) Hg absolute
pressure  prior to testing.  Also, the 16-hour sample residence  time
in  the flask must be observed*   Table  4.1   provides  an  activity
matrix for all on-site activities.

4.  Posttest Operations
    Section3.14.5TPostsampling  Operations)  gives  the  posttest
equipment  procedures  and a step-by-step  analytical  procedure  for
determination of NO , expressed  as N02-  Posttest calibration is not
required  on any of the sampling equipment.  The posttest  operations
form (Figure 5.1, Section 3.14.5) provides  some key parameters to be
checked by the tester  and  laboratory  personnel.   The step-by-step
analytical procedure description  can  be  removed  and  made  into a
separate  quality  assurance  analytical  reference  manual  for  the
laboratory personnel.  Analysis of calibration standards is conducted
in conjunction  with  the  analysis  of  the  field  samples.  Strict
adherence to Method 7A analytical procedures must be observed.

    Section  3.14.6  (Calculations)  provides  the  tester  with  the
required equations, nomenclature,  and  significant  digits.   It  is
suggested  that  a  calculator be used, if available, to  reduce  the
chances of calculation error.

    Section 3.14.7 (Maintenance) provides the tester with a guide for
a maintenance program.  This program  is  not  required,  but  should
reduce equipment malfunctions.   Activity  matrices (Tables 5.1, 6.1,

-------
                                                    Section No.  3.14
                                                    Date July 1,  1986
                                                    Page 7

and  7.1)  summarize all postsampling,  calculation,   and  maintenance
activities.

5.  Auditing Procedure
    Section 3.14.8 (Auditing Procedure)  provides  a  description  of
necessary activities for conducting performance  and  system  audits.
When Method 7A is used  to  demonstrate  compliance with an EPA poll-
utant  emission  standard,  a  performance  audit is required  to  be
conducted of the analytical phase of the method.  The data processing
procedures  and  a checklist for a systems audit are also included in
this section.  Table 8.1 is an activity  matrix  for  conducting  the
performance and system audits.

    Section   3.14.9   (Recommended   Standards    for   Establishing
Traceability) provides the primary standard  to  which  the  analysis
data should be traceable.

6.  References
    Section  3.14.10  contains  the promulgated  Method  7A;  Section
3.14.11  contains  the references  cited  throughout  the  text;  and
Section 3.14.12 contains copies of data forms recommended  for Method
7A.

-------
                                                Section No. 3.14
                                                Date July 1, 1986
                                                Page 8
                     PRETEST SAMPLING CHECKS
                     (Method 7A,  Figure 3.1)

Date 	  Calibrated by 	

Flask Volume
Flask volumes measured with valves?  	 yes  	 no
Volume measured within 10 ml of actual volume?* 	 yes 	 no

Temperature Gauge
Was a pretest temperature correction used? 	 yes 	 no
If yes, temperature correction 	 (within 1°C (2°F)
   of reference values for calibration and within +_ 2°C
   (4°F) of reference values for calibration check).

Vacuum Gauge
Was gauge calibrated against a U-tube mercury manometer (if it
   was a mechanical gauge)?* 	 yes 	 no 	 not applicable

Barometer

Was the pretest field barometer reading within 2.5 mm (0.1 in.) Hg
   of the mercury-in-glass barometer? 	 yes  	 no
*Most significant items/parameters to be checked.

-------
                                               Section No.  3.14
                                               Date  July 1,  1986
                                               Page  9
                      PRETEST PREPARATIONS

                    (Method 7A,  Figure 3.2)
Apparatus check
Probe
Glass liner
clean
Heated properly*
Leak checked
Collection Flask
Clean
Leak checked
Temperature
gauge
Evacuation System
Leak- free pumps
Manifold and
tubing
U-tube manometer
Barometer
Reagents
Water
Absorbing solu-
tion*
Sample Recovery
Dropper or burette
Sample bottles
Pipette, 25-ml
Acceptable
Yes





No





Quantity
required





Ready
Yes





No





Loaded
and packed
Yes





No





*Most significant items/parameters to be checked.

-------
                                                Section No. 3.14
                                                Date July 1, 1986
                                                Page 10
                       ON-SITE MEASUREMENTS
                     (Method 7A, Figure 4.3)
Sampling
Volume of 25 ml of absorbing solution** placed in flask?
Flask valve stopper in purge position? 	
Sampling train properly assembled? 	
  Leak free?* 	  Stopcock grease used?
  Type? 	
Flask evacuated to £75 mm (3 in.) Hg pressure?
  Leakage from manometer observation?*	
  (e.g., maximum change in manometer of £10 mm (0.4 in.)
  Hg/min)     .	
Initial pressure of flask recorded?* 	
Initial temperature of flask recorded? 	
Probe purged before sampling? 	
Sample collected properly?* 	
Flask shaken for 5 min after collection and disassembly from
  train?*	
Samples properly labeled and sealed and stored for shipment?
Sample Recovery
Samples allowed to remain in flasks for minimum of 16 h?*
Final flask temperature and pressure recorded?* 	
Sample transferred to leak-free polyethylene bottle? 	
Flask rinsed twice with 5-ml portions of water and rinse
  added to bottle containing sample? 	
*  Most significant items/parameters to be checked.
** Note that absprbing solution for Method 7A is different from
   that of Method 7.

-------
                                                Section No.  3.14
                                                Date July 1,  1986
                                                Page 11
                       POSTTEST OPERATIONS
                     (Method 7A, Figure 5.1)

Reagents
Sodium nitrate dried at 105° to 110°C for a minimum of 2 hours
  before use? 	
Stock standard solution (sodium nitrate) less than 1 month old?
Sample Preparation
Has liquid level noticeably changed?* 	
  Original volume 	  Corrected volume 	

Analysis
Standard calibration curve prepared?* 	
All calibration points within 7 percent of linear calibration
  curve?*
Reagent blanks made from absorbing solution or eluent solution?

Same injection volume for both standards and samples? 	
Duplicate sample values agree within 5 percent of their mean?
All analytical data recorded on checklist and laboratory form?
* Most significant items/parameters to be checked.

-------
                                                   Section No.  3.14.1
                                                   Date July 1, 1986
                                                   Page 1
1.0  PROCUREMENT OF APPARATUS AND SUPPLIES
    A schematic of the sampling train used for  Method 7A is shown in
Figure 1.1.  The train and sampling procedures are identical to those
for  Method 7.  The sample recovery procedures and equipment are also
identical, with the exception that there is  no  need  to  check  and
adjust  the  pH  of  the  samples.   The  analytical  procedures  and
equipment involved are different.

    Specifications, criteria, and/or  design  features  are  given in
this section to aid in the selection  of  equipment or any components
that are different  from  those  in  Section  3.6.1.   Procedures and
limits  (where  applicable)  for  acceptance  checks  are also given.
Alternative  grab  sampling systems or equipment capable of measuring
sample volume  to within 2% and collecting a sufficient sample volume
to allow analytical repeatability  to  within  5%  may be acceptable,
subject to approval.

    During the procurement of equipment and supplies, it is suggested
that a procurement log be used to record the descriptive title of the
equipment, identification number  (if applicable), and the results of
acceptance  checks.  An example of a procurement log is shown in Fig-
ure  1.2.   A  blank copy of this form is provided in Section 3.14.12
for the convenience of the Handbook user.  Calibration data generated
in the acceptance check are  to  be  recorded  in the calibration log
book.

    The following equipment is that which  is  specified in Method 7A
and has not already  been  described  in  Section 3.6.1 for Method 7.
Table  1.1  at the end of this section summarizes  quality  assurance
activities for the  procurement  and  acceptance of all apparatus and
supplies for Method 7A including the equipment described  in  Section
3.6.1.

1.1  Analysis

    For the analysis, the following equipment is needed.  Alternative
instrumentation  (and  corresponding  procedures)  will  be  allowed,
provided the calibration precision  discussed  in  Section 3.14.2 and
acceptable accuracy can be met.

1.1.1   Volumetric  Pipets - Class-A volumetric pipets are  required.
For making up the calibration  standards,  pipets  of  the  following
sizes are needed: one 1-ml, one 2-ml, one  4-ml,  one  6-ml,  and one
10-ml.   Enough 5-ml pipets  are  needed  for  preparing  calibration
standards, blanks, and samples.

1.1.2  Volumetric Flasks -  Two  Class-A  50-ml volumetric flasks are
needed  for  each  sample,  and one Class-A 50-ml volumetric flask is
needed for each standard  and  each blank.  Also required are Class-A
200-ml  and  Class-A 1000-ml  sizes.   Additional  volumetric  flasks
(50-ml) may be required for audit samples and for dilution of samples
having concentrations in excess of the highest standard.

-------
  PROBE
FILTER
                                FLASK VALVE
                                 FLASK
                            FLASK SHIELOu .\
                                                                            SQUEEZE BULB


                                                                          MP VALVE

                                                                               PUMP
                                              THERMOMETER
            Figure  1.1.   Method  7A evacuated  flask  sampling train.
                                                                                                     •o o w
                                                                                                     0> 0> fl>
                                                                                                     «Q ft O
                                                                                                     
-------
Item description

Qty-

Purchase
order
number
/OSS'
Vendor

Date
Ord.
i/ilesr
Rec.
I/IS/6S
Cost

Disposition
£.& /1 1<>*'&'£&£L '
f/2~ Z'/Q S"
Comments

Figure 1.2.    Example of a procurement log.
                                                                           13 a w
                                                                           0) 0) 0)
                                                                           IQ ft O
                                                                           (D 
-------
                                                   Section No. 3.14.1
                                                   Date July 1, 1986
                                                   Page 4

1.1.3  Analytical Balance - One analytical balance that weighs to 0.1
mg and a set of Class-S calibration weights  to check the accuracy of
the balance (+_ 0.3 mg) upon receipt are needed.  The  balance  should
be serviced or returned to the manufacturer  if  agreement  cannot be
met.

1.1.4  Ion Chromatograph - The ion  chromatograph  should, at a mini-
mum, have the components described below.

    Sample Injection Device - This device must be capable of deliver-
ing a reproducible volume of sample to the ion chromatograph.

    Columns - The ion  chromatograph  should  have an anion separator
column capable of giving duplicate results within 5 percent  of  mean
value and of resolving the  nitrate  ion  from  sulfate  ion and from
other  species  present.   Both  the  Dionex  HPIC-ASC fast run anion
column  for suppressed 1C and the Wescan 269-029 Anion/R  Column  for
non-suppressed  1C  have been demonstrated to give acceptable  separ-
ation.  If suppressed 1C is to be used, an anion suppressor column is
required.  The  Dionex  AFS  anion  fiber suppressor (recommended) or
ASC-1 general purpose suppressor may be used.  Suppressor columns are
generally produced as proprietary items; however,  one can be made in
the  laboratory using the resin available from BioRad  Company,  32nd
and Griffin Streets, Richmond, California.

    Pump  -  The pump must be capable of maintaining a steady  eluent
flow as required by the system.

    Flow Gauges - These must  be  capable  of measuring the specified
eluent flow rate.  It is recommended  that  the  gauge  be calibrated
upon receipt.

    Conductivity Detector with Temperature  Compensation  - It should
be capable of giving responses that can be integrated  with a precis-
ion of  +_  5  percent.   It is recommended that the detector be cali-
brated according to manufacturer's procedures prior to initial use.

    Recorder - It should be compatible with the output voltage of the
detector.

1.2  Reagents

    Unless otherwise indicated,  it  is  intended  that  all reagents
conform  to  the  specifications  established  by  the  Committee  on
Analytical Reagents of the  American  Chemical  Society  (ACS), where
such  specifications  are  available;  otherwise,  use the best grade
available.

1.2.1  Sampling - To  prepare  the absorbing solution, cautiously add
2.8  ml  concentrated  H2S04  to a 100-ml flask containing water (see
specifications in Subsection 1.2.3 below), and dilute  to volume with
mixing.   Add  10  ml  of this  solution,  along  with  6  ml  of  3%

-------
                                                   Section No. 3.14.1
                                                   Date July 1, 1986
                                                   Page 5

hydrogen peroxide  that  has  been freshly prepared from 30% hydrogen
peroxide,  to  a  1-liter  flask.   Dilute  to volume with water (see
Subsection 1.2.3), and mix well.  The absorbing solution must be used
within  1  week of its preparation and, if possible, within 24 hours.
Store in a dark-colored bottle.  Do not  expose  to  extreme  heat or
direct  sunlight.   Refrigerate  the 30% hydrogen peroxide  solution.
Note; The H2S04 content of this absorbing solution  is  10 times less
than that used for Method 7.  The solution is prepared in this manner
to avoid interference from sulfate ions during the analysis by 1C.

1.2.2   Sample  Recovery  -  Use  ASTM  D1193-82, Type III water (see
Subsection  1.2.3)  for  sample  recovery  and   in   making  various
solutions.   At the  option  of  the  analyst,  the  KMnO.  test  for
oxidizable organic matter may be omitted whenever high concentrations
of organic matter are not expected to be present.

1.2.3  Analysis  -  For  the  analysis,  the  following  reagents are
required.

    Water  - Water should be used which conforms with  ASTM   specifi-
cation D1193-82, Type III.  Type III water is prepared  by  distilla-
tion,  ion  exchange,  reverse  osmosis,  or  a combination   thereof,
followed by polishing with a 0.45 ym membrane filter.  The specifica-
tions for Type III water are shown below.

           Specifications for ASTM D1193-82, Type III Water

          Total matter, max., (mg/L)           1.0

          Electrical conductivity, max.,       1.0
            (ymho/cm) at 25 C

          Electrical resistivity, min.,        1.0
            (ymho/cm) at 25

          pH at 25°C                     6.2 to 7.5

          Minimum color retention time         10
            of KMnO., (min)

          Maximum soluble silica, (vg/L)       10

Note;  Mention of "water"  anywhere  in this Section (3.14) refers  to
ASTM  D1193-82,  Type  III water as described above.  By using  water
from the same source for making reagents, calibration standards,  and
eluents  for the ion chromatograph, the effects of trace quantities  of
nitrate  in the water will be negated with regard to sample analysis.
Therefore, a water blank correction is  not necessary in the  develop-
ment of  the calibration curve.

-------
                                                   Section No.  3.14.1
                                                   Date July 1, 1986
                                                   Page 6
    Sodium Nitrate - Dry an adequate amount of sodium nitrate
at 105  to 110 C for a minimum of 2 hours just prior to preparing tRe
standard solution.  (The  analyst should note that potassium nitrate,
KNOo, is used in EPA Method 7; KNO~ is an acceptable alternative  for
Method 7A. )                       J

    Stock  Standard  Solution, 1 mg N0?/ml  -  To  prepare,  dissolve
exactly 1.847 g of dried NaNO3 (or  2.198  g of dried KNCU) in water,
and dilute to 1 liter in a volumetric flask; mix well.  Tnis solution
is stable for 1 month and should not be used beyond this time.

    The  use  of  old  solution  may cause results to be biased high.
Solutions are readily  contaminated  by  microorganisms  that feed on
nitrate ion.  Unquantified  loss  of  nitrate  ion  from the standard
solution causes the high bias.

    Working  Standard Solution, 25 y g NO^/ml - Dilute  5  ml  of  the
standard solution to 200 ml with water in a volumetric flask, and mix
well.

    Eluent  Solution  -  Use an eluent appropriate to the column type
and capable of resolving nitrate ion from  sulfate  and other species
present.   The  following  eluent s  have been  demonstrated  to  give
acceptable separation:

Suppressed  1C — 0.0024M Na2C03/0.003M NaHCXU.   To  prepare,  weigh
1.018  g  of  sodium  carbonate  (Na^COg)   and  1.008  g  of  sodium
bicarbonate (NaHC03), and dissolve in 4 liters of water.

Non-Suppressed  1C  —  0.007M  p-hydroxybenzoic acid,  pH  8.4.   To
prepare, weigh 3.867 g  p-hydroxybenzoic  acid,  and  dissolve  in  4
liters of water.  Adjust to pH 8.4 with lithium hydroxide.

    Quality Assurance Audit Samples - Same as required  by  Method  7
(Section 3.6.8).

-------
                                                              Section No. 3.14.1
                                                              Date July  1,  1986
                                                              Page 7
                 TABLE  1.1.  ACTIVITY MATRIX FOR PROCUREMENT OF APPARATUS
                             AND SUPPLIES

Apparatus/
supplies
Probe
Collection
flask
Flask valve
Temperature
gauge
Vacuum line
tubing
Vacuum gauge
Vacuum pump
Squeeze bulb
Volumetric
pipettes
Acceptance criteria
Borosilicate glass
stainless steel, or Tef-
lon tubing capable of
removing moisture
condensation
Two-liter borosilicate
glass round bottom, short
neck w/24/40 standard
taper opening
Borosilicate glass T-bore
stopcock w/24/40 standard
taper male joint (joint
connection to be made by
glassblower)
Dial- type, capable of
measuring from -5 to
+50°C within 1C
Capable of withstanding
75 mm absolute pressure
U-tube manometer, open
end, 1 m with 1-mm divi-
sions
Pump capable of pulling
vacuum of 75 """ Hg or
less
Rubber, one way
1-, 2-, 4-, 5-, 6-, 10-,
25-ml Class-A glass and
graduated 5~ml
Frequency and method
of measurement
Upon receipt, visually
check for cracks or
flaws and heating capa-
bility
Upon receipt, visually
check, and leak check
Visually check upon
receipt
Visually check upon
receipt, and compare
against Hg-in-glass
thermometer
Upon recer.pt, visually
check and leak check
Visually check upon
receipt
Upon receipt, check with
suitable pressure gauge
Visually check upon
receipt
As above
Action if
requirements
are not met
Return to sup-
plier, and
note in pro-
curement log
As above
As above
As above
As above
As above
As above
As above
As above
(continued)

-------
                                                               Section No. 3-1^.
                                                               Date July 1, 1986
                                                               Page 8
Table 1.1  (continued)
Apparatus/
supplies
Acceptance criteria
Frequency and method
    of measurement
   Action if
 requirements
  are not met
Stopcock
  grease
High vacuum high temper-
ature chlorofluorocarbon
grease
As above
As above
Barometer (or
 consult lo-
 cal weather
 station)
Capable of reading atmos-
pheric pressure to
+2.5 mm Hg
Visually check; cali-
brate against mercury-
in-glass barometer
As above
Storage bottle
Polyethylene, 100-ml, or
greater capacity, screw
cap
Visually check upon
receipt
Return to sup-
plier and note
in procurement
log
Wash bottle
Polyethylene or glass
Visually check label
upon receipt
As above
Analytical
 balance
Capable of measuring
to +0.1 mg
Check with standard
weights upon receipt
and before each use
Replace or
return to man-
ufacturer
Volumetric
 cylinders
50-ml (Class-A) with
1-ml divisions
As above
As above
Ion Chroma-
 tograph
  1. Columns
1. Capable of giving
nitrate ion peaks
with baseline
separation; capable of
giving duplicate results
within 5 percent of mean
value
1. Check during
analyses
1. Consult op-
erator's manu-
al ; regenerate
suppressor
column; clean
separator
column; check
performance
of components
below; replace
column(s) if
above actions
are unsuccess-
ful
 (continued)

-------
Table 1.1  (continued)
                                                              Section No. 3.14.1
                                                              Date July 1, 1986
                                                              Page 9
Apparatus/
supplies
Acceptance criteria
Frequency and method
    of measurement
  Action if
requirements
 are not met
2. Pump
2. Capable of delivering
eluent at constant and
repeatable flow rate
3. Flow
    control
3. Capable of giving
repeatable indications
of eluent flow rate
4. Conduc-
    tivity
    detector
5. Recorder
4. Capable of giving
responses which can be
manually or electron-
ically integrated within
a precision of 5 percent
5. As above, if used to
record responses for
manual integration
2. Check during analyses
by monitoring flow rate
3- Check calibration
and repeatability upon
receipt
4. Calibrate according
to manufacturer's in-
structions prior to use
5. Check during
analyses
2. Consult oper-
ator's manual;
oil, clean, re-
repair, replace,
or return to man-
ufacturer; check
tubing of
ion chroma-
tograph for
leaks or ob-
structions;
check flow meter
performance
3. Consult oper-
ator's manual;
adjust, repair,
replace, or re-
turn to manu-
facturer ;
check pump per-
formance

4. Consult opera-
tor's manual;
Repair, replace,
or return to
manufacturer
5. Consult opera-
tor's manual;
adjust speed
Water
Meets ASTM D1193-82;
Type III
Check each lot, or
specify type when
ordering
Replace, or re-
turn to manu-
facturer
(continued)

-------
Table 1.1  (continued)
                                                               Section No. 3.14.1
                                                               Date July 1, 1986
                                                               Page 10
Apparatus/
supplies
Sulfuric
acid
Hydrogen
peroxide
Sodium nitrate
Sodium carbon-
ate
Sodium bicar-
bonate
p-Hydroxy-
benzoic acid
Acceptance criteria
Concentrated, ACS re-
agent grade
30# aqueous solution,
ACS reagent grade
(store refrigerated)
ACS reagent grade
ACS reagent grade
ACS reagent grade
ACS reagent grade
Frequency and method
of measurement
As above
As above
As above
As above
As above
As above
Action if
requirements
are not met
As above
As above
As above
As above
As above
As above

-------
                                                   Section No.  3.14.2
                                                   Date July 1, 1986
                                                   Page 1
2.0  CALIBRATION OF APPARATUS
    Calibration of apparatus is one of  the  most important functions
in maintaining data quality.  It is highly recommended  that a labor-
atory log book of all calibrations be maintained.  Calibration proce-
dures  for  the  collection  flasks, field  barometer,  thermometers,
vacuum  gauge,  and analytical balance used in Method 7A are the same
as those  described  for  Method  7  (see  Section 3.6.2) and are not
duplicated  in  this  section; a form, however, for use in the analy-
tical  balance calibration is shown in Figure 2.1.  Detailed calibra-
tion procedures  for  the  ion  chromatograph system are described in
this  section.  Table 2.2 at the end of this section  summarizes  the
quality  assurance  activities  for  all calibrations  in  Method  7A
including those described in Section 3.6.2.

2.1  Ion Chromatograph System

    For  Method  7A,  the  calibration of the ion chromatograph  (1C)
system,  except  for  the initial  calibration  of  the  conductivity
detector, is conducted in conjunction  with  analysis  of  the  field
samples.   Specifically,  the  field samples are  analyzed  twice  in
between  three  analyses  of  the ion chromatograph calibration stan-
dards; the exact  sequence  is discussed in detail in Section 3.14.5.
The three analyses of the calibration standards are used to prepare a
calibration curve that is used to determine a calibration  factor for
calculating  the  concentration  of  nitrogen  oxides  in  the  field
samples.  It is, however, highly recommended that the analyst conduct
a preliminary calibration of the 1C any time the system is set up for
analysis  of  NO   field  samples.  For this reason, the full discus-
sion of  the analysis of calibration standards and preparation of the
calibration curve is  presented  in  this section.  Also addressed in
this  section  are  preliminary considerations in  preparing  the  1C
system for use and other considerations for ensuring quality data.

2.1.1  Preliminary Considerations

    Conductivity  Detector - Prior to its initial use, the conductiv-
ity  detector of the ion chromatograph  must  be  calibrated  by  the
method described in the operator's manual.

    Recorder  -  A  strip chart recorder compatible with  the  output
voltage range of the conductivity detector  may be used to record the
ion chromatogram.  Manual measurement techniques that can be used for
quantitation of  the  chromatogram  include (a) peak height, (b) peak
area by triangulation, (c) peak area by multiplying peak height times
the peak width at half-height, (d) peak  area by cutting out the peak
from the chromatogram and weighing it on an analytical  balance,  and
(e) peak area by planimetry.

    The use of an electronic integrator, if available, is recommended
for greater accuracy and precision.  The electronic integrator can be
used in the peak area mode when the integration parameters are set up

-------
                                               Section No. 3.14.2
                                               Date July 1, 1986
                                               Page 2
Balance name
Number
Classification of standard weights
"£'


Date
/Z5&-
0.5000 g
(}. 5£0<4-
1.0000 g

10.000 g

50.0000 g

100.0000 g
/£>O • OOO 4-
Analyst

           Figure 2.1.   Analytical balance calibration form

-------
                                                   Section No.  3.14.2
                                                   Date July 1,  1986
                                                   Page 3

properly.  The key integration parameters for peak area determination
concern the identification of the beginning and end of a peak and the
placement of the baseline under the peak.  Analysts  should carefully
read the operator's manual and understand the selection and set up of
the  integration  parameters  for  their  particular integrator.  The
electronic  integrator  can  also  be  used in the peak  height  mode
provided that the peaks are symmetrical  and  an  acceptable standard
calibration  curve  can be generated without  any  calibration  point
deviating from  the line by more than 7 percent (see Subsection 2.1.3
of this section).

    Sample  Injection  Device  Contamination  Check - The analyst  is
encouraged to check the  sample injection device for contamination by
injecting  water  before  the  calibration  standards  are  analyzed.
Contaminants  will  appear  as  peaks  on the chromatogram.  Repeated
injections of water should be used to remove  contaminants  from  the
sample  injection  device.   If  certain peaks remain  after  several
injections of water then the water may be contaminated and should  be
replaced.

    Separation of Nitrate, NO^  - To ensure accurate results from the
ion chromatographic analysis, baseline  separation of the nitrate ion
(N03~)  peak from the other ion peaks should be achieved.  For_Method
7A, the separation of the N03~  peak from the sulfate ion (S0.~) peak
is of major concern.  The SO4~ originates primarily from the sulfuric
acid  absorbing  reagent.  A second source of S04~ in a sample may be
sulfur dioxide present in the effluent stream sample.   Figures  2.2a
and 2.2b show two chromatograms, one having overlapping N0~~ and SO."
peaks, and the other having baseline separation of the NO.,   and sof=
peaks.  The sulfuric acid concentration in the absorbing reagent used
for Method 7A is 10 times less than that for Method 7 to minimize the
problem of adequately separating N03~ from C>0~.

    The  analyst is encouraged to check the performance  of  the  ion
chromatograph  system before_analyzing samples  in  order  to  ensure
baseline  separation  of N03~ is attainable.   A  test  for  baseline
separation of NO3  can  be  made  by  preparing  a  performance check
sample and analyzing during  the  recommended preliminary calibration
as follows:

    1. Pipet 10.0 ml of the 25 yg N02/ml  working  standard  solution
into a 50-ml volumetric flask.

    2.   Into  the  same  volumetric  flask,  pipet 5 ml of absorbing
reagent.

    3.  Dilute with water to the mark.

    4.   Analyze  this  performance  check  sample  with  calibration
standards  in  the  same  manner  as described for field samples (see
Subsections 5.1.4, 2.1.2, and 2.1.3).

-------
                                     Section No. 3.14.2
                                     Date July 1, 1986
                                     Page 4
                                              so,
Figure 2.2a.  Example chromatogram having
              overlapping peaks.
                                               SO,
                                               NO.
Figure 2.2b.  Example chromatogram showing
              baseline separations of peaks,

-------
                                                   Section No. 3.14.2
                                                   Date July 1, 1986
                                                   Page 5

    The  analyst  should check the chromatogram  of  the  performance
check sample for baseline  separation.  If the baseline separation is
marginal for the performance  check  sample and the samples have N0~~
concentrations close to that of the highest standard  (5  ug N0?/ml7,
the  analyst   should   closely   monitor   subsequent  field  sample
chromatograms  to ensure that results are not adversely  affected  by
deterioration of the ion chromatograph column  or varying performance
of the ion chromatograph.

    The  final aspect of the performance check involves  a  precision
assessment.  The result  from  the  analysis of the performance check
sample should agree within 5 percent of the value for the 5 ug N0~/ml
calibration standard data point.

2.1.2  Preparation of Calibration Standards  - The preparation of the
calibration  standards  is perhaps the most critical  aspect  of  the
Method  7A analysis, since the quality of sample results will only be
as good as the quality of the calibration.  The steps observed in the
preparation of the calibration standards are detailed below.

    Stock Standard Solution

    1.  Dry approximately 5 g ACS-grade sodium nitrate  (NaNO~) in an
        oven at 105  to 110 C for  at  least  2  hours  prior to use.
        Drying  of the NaNOo is necessary to prevent NO  results from
        being biased high because of absorbed moisture.x

    2.  Calibrate the analytical balance using a 2-g Class-S calibra-
        tion  weight  (see  Figure  2.1  for an example  form).   The
        balance reading should agree  within  2  mg  of  the  Class-S
        calibration  weight.  Corrective actions should be  taken  if
        this agreement is not achieved.

    3.  Allow the dried NaN03 to cool to room temperature in a desic-
        cator.   When  the  reagent  has cooled, weigh out 1.847 g to
        +0.002 g.   Cooling  is  required  to prevent weighing errors
        originating from convection  currents.   Storage of the NaNO~
        in the desiccator ensures that moisture will not be adsorbed.

    4.  Place weighed NaN03 in a 1-liter Class-A volumetric flask and
        dissolve in exactly  1  liter  of  water.   Label  the  flask
        accordingly:
                         NaN03(aq)
                         StocR Standard
                         for EPA Method 7A
                         (1 mg N02/ml)
                         Date
                         Analyst's Initials

-------
                                                   Section No. 3.14.2
                                                   Date July 1, 1986
                                                   Page 6

         The solution is stable for one month  and should not be used
         beyond that time.  After about one month, there is increased
         risk that the reagent will be contaminated by microorganisms
         that feed on nitrate.  The use of such contaminated reagents
         will cause NO  results to be biased high.
                      a

    Working Standard Solution

    5.   Pour  about  25  ml of stock standard solution into a clean,
         dry beaker.

    6.   Using a 5-ml Class-A  pipet,  pipet  5  ml of stock standard
         solution into a 200-ml Class-A volumetric  flask.  Dilute to
         the calibration mark with water, and mix well.

         This solution is the Working  Standard;  its nitrate content
         represents  a  concentration  of  25 yg N02/ml.  The working
         standard  solution  is  prepared  fresh  for   each  set  of
         analyses.

    Calibration Standards

    7.   Prepare a series of five calibration standards by  pipetting
         1.0, 2.0, 4.0,  6.0, and 10.0 ml of working standard solution
         (25  yg/ml)  into  a series of five 50-ml Class-A volumetric
         flasks.  The standard masses  will  equal  25, 50, 100, 150,
         and 250 yg N02, respectively. Dilute to the mark with either
         water or eluent solution, and mix well.

         The choice of diluent  is determined by practical considera-
         tions.  If the "water  dip"  (see Figure 2.2) is expected to
         interfere  with  the  nitrate peak of the chromatogram, then
         eluent  should  be used  as  the  diluent  since  this  will
         minimize the  "water dip."  Note: Whichever diluent is used,
         it is important  for the analyst to use. the same diluent for
         the field samples, the calibration standards, and the blank,
         as specified in the Federal Register.

2.1.3   Preparation and Validation of the Calibration Curve -  Method
7A specifies  the  determination of a calibration factor, S, which is
used to calculate the concentration of NO   in  the field samples.  S
is defined as the reciprocal of the slope of the  calibration  curve,
which  is determined by preparing or calculating a linear  regression
plot of the standard  masses of the calibration standards (yg) versus
instrument response (peak height or area).  Determination  of  S does
not take into account the y-intercept, if present, of the calibration
curve.

    The  first subsection that follows describes the calibration pro-
cedures  and the determination of the calibration factor as specified
in Method 7A.  The  second subsection offers an alternative approach,
acceptable  to  the Administrator,  for  conducting  the  calibration

-------
                                                   Section No. 3.14.2
                                                   Date July 1, 1986
                                                   Page 7

calculations that utilize the non-zero y-intercept, if present.  This
approach is based on the calibration  procedures  of  Method  7D  and
involves  the  determination  of a calibration equation.  A data form
which can be used with both approaches is presented in Figure 2.3.

    Determination of the Calibration  Factor  (S) - The determination
of the calibration  factor,  S,  involves  the  three steps presented
below.

    1.  Analyze each of the calibration standards (25, 50, 100,  150,
and 250  yg  N0~)  three times using the ion chromatograph.   Document
chromatograms  fsee  Subsection  5.1.4) and record the results on the
analytical data form for calibration standards (Figure 2.3).  Average
the three responses for each of the five standards.

    2.  Use the average response for  the  five calibration standards
to calculate  the  slope  of  the  calibration curve, graphically, by
least  squares,  or by linear regression.   To  calculate  the  slope
graphically, plot the instrument response (peak height or area count)
on  the  y-axis against the corresponding NG>2 standard  concentration
value on the x-axis.  Draw a "best-fit"  line  between the points and
determine the slope of the line.  Least squares (a method  acceptable
to the Administrator) can be  hand  calculated and is shown in Figure
2.3.   To  calculate the slope by  linear  regression,  use  the  NO.-
standards as the independent variable (x-axis)  and the corresponding
instrument response as the dependent variable (y-axis).

    3.  The calibration factor, S, is calculated as the reciprocal of
the slope of the calibration  curve,  determined  from the "best-fit"
line or the linear regression equation.  Any y-intercept is ignored.

    4.  The calibration factor, S, and  therefore,  the curve must be
validated.  Using the calibration  factor  for  calculation, the pre-
dicted sample mass for each calibration standard is compared with the
known  value for that standard.  The predicted sample mass  must  not
deviate from the known standard concentration  by  more than 7%.  The
quantity  "yg  N02  Predicted"  is  calculated  using the calibration
factor (S) and the detector response  (H), in millimeters or integra-
tor response, as shown in Equation 2-1.


                                                         Equation 2-1
    yg N02   = S (yg/mm) x Detector (mm)
    Predicted              Response
                              H

The deviation of each predicted  sample  mass  from the known mass is
calculated using Equation 2-2.
                                                         Equation 2-2

    Deviation = ^-^2 Predicted - yg N02 Standard x 1QQ%
      (%)                    y g N02 Standard

-------
Plant

Date
          / /Z
                            Location

                            Analyst .
                                                              Section No. 3.14.2
                                                              Date July 1, 1986
                                                              Page 8
 Was an integrator used?
                            yes
Was the intercept (I) used for calculations? yes S ho
Were all points within 7 percent of calculated value? */ yes
Sample
Identifier
Std I
Std 2
Std 3
Std 4
Std 5
Sample
Mass
(yg NO,)
25
50
100
150
250
Integrator Response
or Peak Height (mm)
H
1
4,1
/ZS"
2.5*3
381
6>0.\
2
6.4-
IZ-1
Z+3
39. /
5W
3
6.2
/3-0
zs.r
30.fr
57. <*
Avg
6-23;
IZM
ZS".23
3'0. (r0
61, &t
Predicted
Sample Mass
(yg NO,)

52. //
/^Z, 72-
/r/./r
>^f-3, 7r
no
Deviation
/ — / ^ /
t * ^f^f
+ 4.23
^™ ^ ^*^
+-4.77
-^.5Z>
Predicted Sample Mass using Least Squares to Calculate Calibration Factor (S)
  with Zero Intercept
S = S1H1
S2H2
                     S3H3
                                    S5H5
        H
                      (/2.0Q)
   S = 4:07/3  y g N02/mm

  Predicted Sample Mass (yg N02)

   yg N0_ = H x S = (^.Z3 ) x
                                                            Equation 2-1
Predicted Sample Mass using Linear Regression to Calculate Calibration Factor (S)
  and Non-Zero Intercept (I)
    y = mx + b; m =
    x = - (y - b); - = S =
                            .5 b
                             1
        m
                   m
    y = H; and b = I (Intercept) =

   Predicted Sample Mass (yg N0_)

    yg N02 = S(H - I)

    yg N02 at 25 yg standard = 	
                                                          Equation 2-
        Figure 2.3.  Analytical data form for analysis of calibration standards.

-------
                                                   Section No.  3.14.2
                                                   Date July 1, 1986
                                                   Page 9

This calculation is performed for each calibration standard using the
average of the three response measurements.  If any point (known con-
centration of standard) deviates from the line (predicted  concentra-
tion)  by more than +7 percent, that standard should  be  remade  and
reanalyzed.

    Linear regression using a hand-held calculator  is recommended to
obtain the slope (and equation) for the calibration  curve.  Inexpen-
sive calculators  are available which have linear regression programs
that are quick and simple to use.  Graphical techniques are relative-
ly simple matters when  all  the  calibration  data  points lie on or
close to the line.  However,  when  deviations  from linearity occur,
the  placement  of the "best-fit" line becomes ambiguous because  the
data points are not evenly distributed.

    Determination  of  the  Calibration  Equation - As discussed pre-
viously, Method 7A directs that the calibration factor, S, be used to
calculate the field sample analytical results.   in  cases  where the
calibration curve does not pass through the origin, the procedure  of
Method  7A  could give biased results for both the field samples  and
the linearity check since the equation for the calibration curve will
contain an intercept term not taken into account in the calculations.
Accordingly,  this section offers an alternative calibration approach
adapted from  Method  7D.   The  approach involves determination of a
calibration  equation  which takes into account both the slope of the
calibration  curve  and  any  y-intercept  term  and which is used in
calculating the NO  concentration of field samples.
                  H

    Derive  the linear calibration equation  or  curve  using  linear
regression.   The  calibration  equation  should  be expressed in the
following form:

          y = m x + b                                    Equation 2-3

where

         m = slope of the linear calibration curve, which is equal to
             the reciprocal of the calibration factor, 1/S, and

         b = y-intercept of linear calibration curve  which  will  be
             referred to as "I" for purposes of later calculations.

   As discussed in the previous section, Method 7A requires that none
of the calibration data points deviate from the calibration  curve by
more than 7 percent  of  the  concentration at that point.  Method 7A
(Section  5.2.3)   states   that  deviations  can  be  determined  by
multiplying  the  calibration factor S times the peak height response
for each  standard.   When the calibration equation with intercept is
used, the quantity " yg NO, Predicted" is computed using the following
equation:

-------
                                                   Section No.  3.14.2
                                                   Date July 1, 1986 •
                                                   Page 10

   . yg N02   = S (yg/mm) /Detector (mm) - I (mm)\          Equation 2-4
  Predicted             I  Response             I
                        \    H                  /

As   before,  calculation  of  the  %  deviation  from  the  line  is
accomplished using Equation 2-2.  If any  deviation  is  greater than
7%,  the corresponding standard should be remade and reanalyzed.   If
this does  not  result  in  improved  results,  other  approaches are
discussed in the following subsection "Other Considerations."

2.1.4  Other Considerations - Method 7A requires that if any calibra-
tion standard  point  deviates from the standard calibration curve by
more than 7%, then that corresponding calibration standard is  to  be
remade and reanalyzed.  This corrective action may not always  reduce
the calibration point deviations below 7%.  Some potential causes for
deviation  of  the  calibration  points  from  the calibration  curve
include (a) improper pipetting procedures used to prepare calibration
standards, (b) improper technique for manual  sample  injection  into
the  ion  chromatograph,  (c)  inaccurate  measurement   of  the  ion
chromatograph response, and (d) non-linear  detector response.   Table
2.1 shows the precisions for calibration operations for Method 7A.


                  TABLE 2.1.  TARGET PRECISIONS FOR
                 CALIBRATION OPERATIONS OF METHOD 7A
   	Operation	Precision Target (%)

    Pipetting                                    1

    Introduction of Samples                     <1
       into Ion Chromatograph

    Measurement Response
       o Peak Height                            1-4
       o Triangulation                           4
       o Height X Width at
           Half-Height                           3
       o Electronic Integration                <0.5
    Pipetting  Procedure  and Pipetting Errors  -  In  preparing  the
calibration standards,  pipetting is the most critical step.  Serious
errors  can  originate from poor pipetting  technique.   In  general,
errors will appear as high biased NO  results.  The correct pipetting
procedure is described below.       x

-------
                                                   Section No. 3.14.2
                                                   Date July 1, 1986
                                                   Page 11

    The  pipet should be inspected before use and checked  to  ensure
that the tip is not chipped.   The pipet should be replaced if a chip
is observed.

    The pipet should be rinsed  with  the  reagent to be pipetted and
checked for cleanliness before use as follows.  Approximately 2 ml of
reagent is drawn into  the pipet, which is then rotated and tilted in
order to expose the inner  surface  to the solution.  The rinse solu-
tion is then allowed to drain freely  from  the  pipet  into a beaker
assigned for waste.  If the pipet is clean, the analyst will observe,
after about 10 seconds, that all the rinse solution will have drained
from the pipet with the exception of a  small  quantity  remaining in
the  tip.   If  this  is  not  observed, either the pipet  should  be
cleaned, or another pipet should be obtained.   The  rinse  and check
for cleanliness should be performed at least once.

    For the actual pipetting, reagent  is  drawn into the pipet until
the liquid meniscus is above the calibration mark.  The pipet is then
withdrawn from the solution and the end  is  wiped  with a laboratory
tissue.  Next, the pipet is brought to  a  vertical  position and its
tip  is brought to touch the inside of the beaker assigned for waste.
The liquid  in  the  pipet  is then allowed to drain slowly until the
meniscus coincides with the calibration mark.

    The pipet is then transferred to  the  appropriate container and,
with the pipet in a vertical position and its tip touching the inside
wall of the container, the liquid is allowed to drain freely  into the
container.  The pipet's tip should be kept  in  contact with  the wall
for  roughly 10 seconds after the liquid has apparently drained.  The
pipet is then removed from the container without disturbing the small
amount of liquid remaining in the tip.

    It is important to recognize that  Class-A  pipets are calibrated
in a manner  which  accounts  for  the  drainage  time and the liquid
remaining in the tip.   If  dirty  pipets  are  used or if the proper
draining  technique is not observed, NO  results will be biased high.
Low biases will occur  if  the   liquid xremaining in the pipet tip is
blown  out  into  the receiving  container.  The significance  of these
biases depends on the size of the  pipet  involved.  For example, the
error with a dirty 25-ml pipet may be  undetectable,  while the error
for a 1-ml pipet can easily exceed 10 percent.

    The precision of the pipetting operation can be checked gravimet-
rically using water.  The technique involves pipetting a known volume
of water into a tared  container  and  determining  the weight of the
water.  The precision  of  the   pipetting operation is estimated from
the results of several repetitions.

    The procedure for manually injecting a sample into the ion chrom-
atograph can be a source  of  error  when analyzing calibration stan-
dards, field samples,  and  QA   samples.   For  fixed  loop injection
systems,.considerable variation can result  from injecting the sample

-------
                                                   Section No. 3.14.2.
                                                   Date July 1, 1986
                                                   Page 12

into  the  loop too fast, resulting in  the  sample  loop  not  being
completely filled.  A slow,  deliberate  injection of the sample into
the  loop  will  completely  fill  the  loop.  The precision  of  the
injection procedure can be checked by performing repetitive  analyses
on a single sample.

    Chromatogram  Quantitation  -  The choice of quantitation methods
for the ion chromatograms can also be a source  of  error when analy-
zing calibration standards,  field samples, and QA samples.  As shown
in Table  2.1, measurement of the detector response by manual methods
has a higher degree  of  imprecision compared to measurement by elec-
tronic  integration.   Method  7A states that peak height measurement
can be used provided  the  peaks  are symmetrical and the required 7%
deviation of calibration points from the standard  calibration curves
can be met.  The peak  height measurement method, even with symmetri-
cal peaks, may not produce a linear standard caliration curve because
the  peak width of the higher concentration standards will  typically
be wider than  the  peak  width of the lower concentration standards.
Figure 2.4 shows the difference  in  the  linearity  of ion chromato-
graphic  calibration  curves  using  the peak area mode and the  peak
height mode.   The  dead  volume  of  the  ion  chromatograph system,
particularly suppressed ion chromatograph systems,  can  also  affect
the peak width.  Quantitation by peak area measurement will eliminate
the   biases  caused  by  widening  peaks  provided  the  peak   area
measurement is done properly.  The use of an electronic integrator in
the peak area mode for ion chromatograms  with baseline separation of
the nitrate peak will produce the most precise calibration curves and
subsequent accurate analyses of field samples and QA samples.

-------
Response
                                                Section No. 3.14.2
                                                Date July l, 1935
                                                Page 13
                       Peak Area
                       Approach
                                  Peak Height
                                  Approach
                               150
2,00
2.50    300
  100
 P g N02

Linear and non-linear ion chroma-
tographic calibration curves.

-------
                                                              Section No.  3.14.2
                                                              Date July 1, 1986
                                                              Page 14
               TABLE 2.2.  ACTIVITY MATRIX FOR CALIBRATION OF EQUIPMENT
Apparatus
Acceptance limits
Frequency and method
    of measurement
  Action if
requirements
 are not met
Collection
  flask
Measure volume within
10 ml
On receipt, measure with
graduated cylinder
Recalibrate
Barometer
Reading agrees within
2.5 mm (0.1 in.) Hg of
mercury-in-glass
barometer
Upon receipt and before
each field test
Repair
or return
Thermometer
Reading agrees within
1°C (2 F) of mercury-
in-glass thermometer
As above
As above
Vacuum gauge
  (mechanical
  only)
Reading agrees within
2.5 mm (0.1 in.) Hg
of mercury U-tube mano-
meter
As above
As above
Analytical
  balance
Weight within 2 mg of
standard weights
(Class S)
Use standard weight
before preparation of
working solution
Repair or
return to
manufacturer
Ion chroma-
  tograph
Calibration curve should
be linear; data points
for calibration stan-
dards must not deviate
from the linear calibra-
tion curve by more than
+7 percent
With each set of field
samples; calibration
standards prepared from
sodium nitrate
Interpret data
using another
technique:  e.g
if using peak
height, change
to peak area;
conduct addi-
tional analy-
ses of cali-
bration stan-
dards; cali-
brate conduc-
tivity detec-
tor; consult
operator's
manual

-------
                                                   Section No. 3.14.3
                                                   Date July 1, 1986
                                                   Page 1
3.0  PRESAMPLING OPERATIONS
    This section addresses  the  preparation  and packing of supplies
and equipment needed for the sampling.  The pretest  preparation form
(Figure 3.1) can be used as an equipment checklist.  Many presampling
operations for Method  7A  are identical to those for Method 7.  This
section will only discuss  the operations that are different; however
all quality assurance activities for Method 7A presampling operations
are  summarized in Table 3.1 at the end  of  this  section  including
those  described in Section 3.6.3.  See Section 3.0 of this  Handbook
for details on preliminary site visits.

3.1  Apparatus Check and Calibration

    Previously used equipment should be  visually  checked for damage
and/or excessive  wear  before  each  field  test.   Items  should be
repaired  or replaced (as applicable) if judged to be unsuitable  for
use.  A pretest  sampling  checks form (Figure 3.1) summarizes equip-
ment calibration.   The pretest preparations form (Figure 3.2) can be
used as an equipment check  and  packing  list.   The  completed form
should be dated, signed by  the  field  crew supervisor, and filed in
the operational log book.  The replacement  of  worn or damaged items
of equipment should  be  initiated.   Procedures  for  performing the
checks are given herein; a check is placed  in  the  proper  row  and
column  as the check/operation is completed.  Each team will have  to
construct its own checklist according to the  type  of sampling train
and equipment it uses.

3.2  Reagents

    Unless otherwise indicated,  it  is  intended  that  all reagents
conform  to  the  specifications  established  by  the  Committee  on
Analytical Reagents of the  American  Chemical  Society  (ACS), where
such specifications  are available; otherwise, use the best available
grade.  See Subsection 1.2.3 of Section 3.14.1 for  water  specifica-
tions.

3.2.1   Sampling - The absorbing reagent is prepared by adding 2.8 ml
of concentrated sulfuric acid (H^SO.)  to  a  100-ml flask containing
water  and  diluting to volume  witn  mixing.   Add  10  ml  of  this
solution,  along  with  6  ml of 3% hydrogen peroxide (H202) that has
been  freshly  prepared  from  a 30 percent solution,  to  a   1-liter
flask.   Dilute  to  volume  with water, and mix well.  Prepare fresh
absorbing solution weekly, and avoid exposure  to  extreme heat or to
direct  sunlight,  as  these  will  cause  the hydrogen  peroxide  to
decompose.  If the reagent  must  be  shipped  to  the  field,  it is
advisable that the absorbing reagent be prepared fresh on-site.

3.2.2   Analysis - The following reagents are needed for analysis and
standardization:

-------
                                               Section No. 3.14.3
                                               Date July 1, 1986
                                               Page 2
Date    ^/<£5 /0S	 Calibrated by	           	
         r    "^
Flask Volume
Flask volume measured with valves?     
-------
                                             Section No.  3.14.3
                                             Date July 1,  1986
                                             Page 3


Apparatus check
Probe
Glass liner
clean
Heated properly*
Leak checked
Collection Flask
Clean
Leak checked
Temperature
gauge
Evacuation System
Leak- free pumps
Manifold and
tubing
U-tube manometer
Barometer
Reagents
Water
Absorbing solu-
tion*
Sample Recovery
Dropper or burette
Sample bottles
Pipette, 25 ml

Acceptable
Yes

/
^
X
x

X
X



X
X

x
X

X
^

X
X
X
No
























Quantity
required

3



14
T




z


2-
1

/ likr
1 lifer

2
14-
1 T
2-

Ready
Yes

X



X





<^
,

iX
X

X
X

^
^
^
No























Loaded
and packed
Yes

i^-"



^





•/
X

X
x

x
X

^^
^s"^
X
No























*Most significant items/parameters to be checked,
             Figure 3.2.   Pretest preparations.

-------
                                                   Section No. 3.14.3
                                                   Date July 1, 1986
                                                   Page 4

    Stock  standard  solution  - Dissolve exactly 1.847  g  of  dried
sodium nitrate (NaN03) [or 2.198 g of dried potassium nitrate (KNOq)]
in  water,  and dilute to 1 liter in a volumetric  flask;  mix  well.
Prepare fresh after 1 month.

    Working standard solution -  Dilute 5 ml of the standard solution
to 200 ml with water in a volumetric flask, and mix well.   Note; One
ml  of  the working standard  solution  is  equivalent  to  25  g  of
nitrogen dioxide.

    Eluent solution - Weigh  1.018 g  of sodium carbonate (NaCO,~) and
1.008 g of sodium  bicarbonate  (NaHCOg), and dissolve in 4 liters of
water.  Other eluents may be used (see Subsection 1.4.3).

-------
                                                               Section No. 3.14.3
                                                               Date July 1, 1986
                                                               Page 5
               TABLE 3.1.  ACTIVITY MATRIX FOR PRESAMPLING PREPARATION
Characteristic
Acceptance limits
Frequency and method
    of measurement
  Action if
requirements
 are not met
Apparatus Check

Probe
1. Clean; glass liner
inert to oxides of
nitrogen

2. Heating properly if
equipped with heating
system

3. Leak free
1.   Before each test
                                           2.  As above
                                           3.  Pressure <380 mm
                                           (15 in.) Hg
Replace
                          Replace or
                          repair
                          Replace or
                          repair
Collection
  flask
Clean; volume within
10 ml
Before each test,
clean with strong de-
tergent and hot tap
water, and rinse with
tap water and then
ASTM Type III water;
periodically clean
with grease remover
Repeat cleans-
ing of flask
and/or meas-
ure volume
Evacuation
  system
Vacuum of 75 nun
(3 in.) Hg absolute
pressure in each flask;
leakage rate <10 mm
(0.4 in.) Hg/min
Before each test, check
for leaks using Hg-
filled U-tube manometer
Correct leaks
Absorbing
Reagent

Sulfuric acid
  concentrated
Final concentration:
0.28 ml/liter
Prepare fresh absorbing
solution weekly; use
graduated pipette
Make up new
solution
Hydrogen perox-
  ide, 3%
6 ml/liter
Water
Deionized distilled
to ASTM specifications
D 1193-82, Type III
                          Prepare fresh
                          for each anal-
                          ysis period
(continued)

-------
TABLE 3.1.  (continued)
                                                                Section No.  3-14.3
                                                                Date July 1, 1986
                                                                Page 6
Characteristic
Acceptance limits
Frequency and method
    of measurement
  Action if
requirements
 are not met
Analytical
Reagents

Stock standard
  solution
1. 1.84? +0.001 g
NaNO- ACS reagent
grade into a 1-liter
volumetric flask
(Class-A)

2. Stored for less
than 1 month
1. On makeup of solution
  use analytical balance
                                           2. Date solution
1. Make up new
  solution
                          2. As above
Working standard
  solution
5 ml of stock solution
into 200-ml volumetric
flask (Class-A)
On makeup of solution,
use Class A pipet and
proper technique
As above
Eluent solution
1.018 g + 0.001 g of
NaCO- and 1.008 g
+ 0.002 g of NaHCO,
in 4 liters       5
On makeup of solution,
  use analytical balance
As above
Packing Equip-
ment for Ship-
ment

Probe
Rigid container lined
with polyethylene foam
Prior to each shipment
Repack
Collection
  flasks and
  valves
Rigid container lined
with polyethylene foam
As above
As above
Evacuation
  system, tem-
  perature
  gauges,
  vacuum lines,
  and reagents
Sturdy case lined with
polyethylene foam
As above
As above
Evacuation
  pump
Shipping container
or housing designed
for travel
As above
As above

-------
Plant fTCsKji rorier r/^nt
Sample location £Sr 0<>i~H£--l> y Bretf-e^ & 1
Operator &&C?

City
Date
C,O^/f^^>a^ ^ Jvfcr>ir-s>Hf*
2/Z7/&ST
Barometric pressure (PK,^>- 2?- 84~




in. Hg



Sample
number
M-\
M-Z


Sample
point
location
6-l\
Jj>~ /O
CIO

Sample
time
24-hr
£733
074S-


Probe
temperature ,
°F
•Z./0
2-/0


Flask
and valve
. number
f-/3
Be -to

Volume
of flask
and valve (V^) ,
ml
2-0/3
Zo/O
r
Initial pressure
in. Hg

Leg A±
/3.6
13.7


Leg B±-
/3V
/3.8 '
/3V.

P ^
2^ . 5**?*
Z.3f
2.4^

Initial temperature

^(ti)
73
73


°R(Ti)
^53
5-35

 Pi = pbar -  (D
                                                                                    
-------
                                                   Section No.  3.14.4
                                                   Date July 1,  1986
                                                   Page 1
4.0  ON-SITE MEASUREMENTS
    The on-site activities include transporting equipment to the test
site, unpacking and assembling  the  equipment, confirming duct meas-
urements  and  traverse  points  (if  volumetric  flow  rate is to be
determined),  determining  the  molecular  weight  of  the stack gas,
sampling  for  oxides  of  nitrogen,  and  recording the data.  These
activities are the  same  as  for  Method 7 (Section 3.6.4), with the
exception of a portion of the sample recovery procedures as described
below.  Blank  data  forms  can  be found in Section  3.14.12 for the
convenience  of  the  Handbook user.  Table 4.1 at the  end  of  this
section  summarizes  the quality assurance activities relative to all
on-site  measurements  in  Method  7A,  including those described  in
Section 3.6.12.

4.1  Sampling

    On-site sampling procedures for Method  7A  are the same as those
for  Method 7.  See Subsection 4.3  of  Section  3.6.4  for  detailed
descriptions of sampling  procedures.   For  convenience, examples of
completed field data  forms  for  Method  7  are  reproduced  in this
section (Figures 4.1A and 4.IB); blank copies are provided in Section
3.14.12.

4.2  Sample Recovery

    Sample recovery procedures should  be  performed as described for
Method 7 (Section 3.6.4),  with  the  exception  that  the  steps for
checking  and  adjusting the pH of the sample should be deleted (note
changes in Figures 4.2A, 4.2B, and 4.3).

    A 16-hour minimum  sample  absorption  period  is  required as in
Method 7.  Samples should be recovered within 4 days of collection.
As currently written,  the  method  states that the samples should be
stored no more.than 4 days between collection and analysis.  However,
a recent study   utilizing samples from nitric acid  plants and power
plants  indicates  that  the  storage  period  between  recovery  and
collection may be extended to 30 days.

-------
   Plant
   Sample  location

   Operator
City

Date
Barometric pressure  (Pbar)
                                               mm Hg


Sample
number
^-/
M>-t
Af>-3

Sample
point
location
*-'/
&-10
c-io

Sample
time
24-hr
0733
074-^
O&oi

Probe
temperature ,
°C
100
loo
loo

Flask
and valve
number
£%r&
K-/6
££-<&
Volume
of flask
and valve (Vp) ,
ml
2013,
ZO/0
ZOO'd
Initial pressure
in. Hg

Le.g Ai
37Z
373
57^.5-

Leg Bi
371
370.^
370

^ia
17. i
/(*.7
n.7

Initial temperature

°C (t±)
Z2. 2
zi.z.
^3.r

°R(Ti)b
2fs:z
2-14. 7.
Z?6.S~
Pi = pbar
                                                                                                               •O D W
                                                                                                               0) 0) (D
                                                                                                               U3 ft O
                                                                                                               0> (D ft
                                                                                                                   H-
                                                                                                               W LI O
                              Figure 4.IB.  Nitrogen oxide  field data form (metric units)
                                                                                                                   O
                                                                                                                   *


                                                                                                                   CO
                                                                                                                 00
                                                                                                                 en

-------
     Plant
                    rs
                      s*Je-r
                                               Date
    Sample recovery personnel  £?. O(c&r    Barometric pressure, (P,   )


    Person with direct  responsibility for recovered samples   /y. £ .
                                                                                       in.  Hg


Sample
number
A/M
frp-Z.

Final pressure,
in. Hg

Leg Af
/fr
1.2
2.0

Leg Bf
o.(.
o.$
/.o

pfa
2-1 M
J27£f
J$~-W

Final temperature,

°F (tf)
73
72-
73

°R (Tf)b
£33
S"32

Sample
recovery
time,
24-h
/3Z2.
1 $4-o


Liquid
level
marked
^
^

Samples
stored
in locked
container
^
LiX*"^

P  — D    «/A  -4- T)  ^         **P  ss *•
pf ~ pbar    (Af + Bf'*         xf   fcf
                                        460°F.
Lab person with  direct responsibility for recovered samples

Date recovered samples received   3// IvS      Analyst   £'.

All samples  identifiable?

Remarks
                                          All liquids at marked level?
                               I
 Signature of  lab sample  trustee
    Figure 4.2A.   NOX sample recovery and integrity data form   (English units)
                                                                                                   V O W
                                                                                                   tu 0) (D
                                                                                                   «Q rt O
                                                                                                   (D (D ft
                                                                                                    M •
                                                                                                    -*

                                                                                                      GO
                                                                                                    l-i •

                                                                                                    VD t->
                                                                                                    00 it^

-------
    Plant
                      ro
                                              Date
    Sample recovery personnel
                                              Barometric pressure, (Pbar)
mm Hg
    Person with direct  responsibility for recovered samples


Sample
number
AA/
A^'Z
AP-3
Final pressure,
' mm Hg

Leg Af
4*.(,
?>o.<£~
5Z?.8

Leg B£
/^.Z
^0.3


pfa
702
7^>7


Final temperature,

°C (tf)
22.?
2Z.Z
22. 7-

°K (Tf)b
MS.?
2<7£~.2

Sample
recovery
time,
24-h
1322
1330
131-1

Liquid
level
marked
•x
^

Samples
stored
in locked
container
^
^

     bar
                                 =  t
                                        273°C.
Lab person with  direct  responsibility for recovered samples

Date recovered samples  received   ^  //^S""   Analyst  £•".

All samples identifiable?

Remarks
                                        All liquids at marked level?
Signature of lab sample  trustee
                                        "P.
   Figure 4.2B.   NO  sample recovery and  integrity data form (metric units)
            •d a w
            o> 0) a>
           tQ ct O
            a> a> ft
                H-
            01 Q O
              C 3
                                                                                                     GJ
                                                                                                   VO
                                                                                                   00

-------
                                               Section No. 3.14.4
                                               Date July 1, 1986
                                               Page 6
Sampling
Volume of 25 ml of absorbing solution** placed in flask?
Flask valve stopper in purge position?
Sampling train properly assembled?
  Leak free?* 	\/       Stopcock grease used?
  Type? 	C,
Flask evacuated to 75 nun (3 in. ) Hg pressure?
  Leakage from manometer observation?* _ ^ / //?.
  [e.g., maximum change in manometer of £ 10 mm (0.4 in.)
  Hg/min]
Initial pressure of flask recorded?*
Initial temperature of flask recorded?* _
Probe purged before sampling? _ S
Sample collected properly?* _ •_
Flask shaken for 5 min after collection and disassembly
  from train?*
Samples properly labeled and sealed and stored for shipment?
Sample Recovery
Samples allowed to remain in flasks for minimum of 16 h?*
Final flask temperature and pressure recorded?*
Sample transferred to leak-free polyethylene bottle?    >^
Flask rinsed twice with 5-ml portions of water, and rinse
  added to bottle containing sample? _ t^
 * Most significant items/parameters to be checked.
** Note that absorbing solution for Method 7A is different from
    that of Method 7.
                Figure 4.3.   On-site measurements.

-------
                                                                Section No. 3.14.
                                                                Date July 1, 1986
                                                                Page 7
                Table 4.1.  ACTIVITY MATRIX FOR ON-SITE MEASUREMENTS
Characteristic
Acceptance limits
Frequency and method
    of measurement
  Action if
requirements
 are not met
Apparatus
 assembly
Assemble using Fig.
1.1; no leakage
Before sample collec-
tion, visually and
physically inspect all
connections
Check for
leaks; repair
system;
repeat test
Operational
 check
Maximum vacuum of
75 mm (3 in.) Hg
absolute pressure
                   Leakage rate £10 mm
                   (0.4 in.) Hg/min
Before sample collec-
tion, use Hg-filled
U-tube manometer
                        As above
Check system
for leaks;
check vacuum
pump

Check all
joints and
valves for
source of leak
Sample
 recovery
Shake flask for 5 min

Let flask set for a
minimum of 16 h, but
no more than 4 days
                   Shake  flask  for 2 min

                   Determine  flask pres-
                   sure and temperature

                   Mark sample  level on
                   container

                   Record data  on data
                   form (Fig. 4.2)
                                           During each sample
                                           collection, use mano-
                                           meter and Celsius
                                           thermometer
                          Reject sample,
                          rerun test
Sample logistics
Properly label all
containers, etc.

Record all data on
field data forms
Visually check each
sample

As above
Complete the
labeling

Complete the
data records

-------
                                                   Section No.  3.14.5
                                                   Date July 1, 1986
                                                   Page 1
5.0  POSTSAMPLING OPERATIONS
    The postsampling operations include  checks  on (a) the apparatus
used  in  the  field to quantify sample volumes (volume, temperature,
and pressure measurements), and (b) analyses of the samples collected
and forwarded to the base laboratory.  If the laboratory receives the
samples in the sample flasks, laboratory personnel will  have to com-
plete the sample recovery procedures referred to in Section 3.6.4.

    The  postsampling  checks on the sample collection train are  the
same as for Method 7 (Section 3.6.5).  The analytical  procedures for
Method  7A  are  different from Method 7  and  are  discussed  below.
Figure 5.1 is a checklist for  all  Method  7A  posttest  operations.
Table 5.1 at the end of this section summarizes the quality assurance
activities  for  all  postsampling operations for Method 7A including
those described in Section 3.6.5.

5.1  Analysis (Base Laboratory)

    Calibration  of  the ion chromatograph, including preparation  of
the calibration standards and preparation of the field  samples is of
primary importance to a precise  and  accurate  analysis.  For Method
7A, the calibration of the 1C is conducted in conjunction with analy-
sis  of  the  field  samples  (and quality assurance samples).   This
section  presents  the  steps  for  analysis  of  the  field  samples
including preparation of samples, field blanks,  and  use  of quality
assurance samples.  The relationship  between  analysis  of the field
samples and  preparation  of  the  calibration  curve  is  addressed.
However, because a  calibration and performance check of the 1C prior
to conducting any NO  analyses is highly  recommended,  the  detailed
discussion  of  the TC calibration is presented  in  Section  3.14.2.
Therefore, the analyst should  use Section 3.14.2 in association with
this section (3.14.5) in conducting the analysis.  In particular, the
analyst  is encouraged to review the discussion of  pipetting  errors
(see  Subsection  2.1.4).   Upon  completion  of  each  step  of  the
preparation of the calibration curve and of each sample analysis, the
data should be entered on the proper data form.

5.1.1  Preparation of Field Samples - Check the  level  of the liquid
in  the  sample  container  and  confirm  whether any sample was lost
during shipment; note this  on  a  data  form  such  as that shown in
Figure 5.1.  If  a  noticeable amount of leakage has occurred, either
void  the  sample  or  use  methods  subject  to the approval of  the
Administrator  to correct  the  final  results.   Immediately  before
analysis  prepare  each  field  sample.   The  following steps detail
sample preparation operations.

    1.  With the aid of a funnel, transfer  the contents of the samp-
        ling flask to a 50-ml Class-A volumetric flask.

    2.  Add approximately  a  5-ml  portion  of water to the sampling
        flask,   replace the stopcock (ensuring  that  it  is  in  the

-------
                                               Section No. 3.14.5
                                               Date July 1, 1986
                                               Page 2
Reagents
Sodium nitrate dried at 105° to 110°C for a minimum of
  2 hours before use?
Stock standard solution (sodium nitrate) less than 1 month old?
Sample Preparation
Has liquid level noticeably changed?* 	/Vp	
  Original volume 	  Corrected volume 	

Analysis
Standard calibration curve prepared?* 	•_	
All calibration points within 7 percent of linear calibration
  curve?*
Reagent blanks made from absorbing solution? 	^
Same injection volume for both standards and samples? 	
Duplicate sample values agree within 5 percent of their mean?

All analytical data recorded on checklist and laboratory form?
* Most significant items/parameters to be checked.
                Figure 5.1.  Posttest operations.

-------
                                                   Section No.  3.14.5
                                                   Date July 1, 1986
                                                   Page 3

       closed  position),   and rinse  the  interior  by  shaking  and
       rotating  the  flask.   Transfer  the rinse to the  volumetric
       flask.  Repeat the rinse with another  5-ml  portion of water,
       and add this rinse to the volumetric flask also.

    3.  Reassemble the sampling  flask  and  place the stopcock in the
       closed  position to prevent contamination during storage prior
       to reuse.

    4.  Using water, dilute the contents  of  the  volumetric flask to
       the mark.  Mix the contents of the flask well.

    5.  Using a 5-ml Class-A pipet, pipet a 5-ml aliquot of the sample
       into another 50-ml  Class-A volumetric flask.  This aliquot is
       diluted to the mark with either water or eluent solution.  Mix
       the contents of the flask well.

    The diluent used must be the same  as  that used for the calibra-
tion standards.  (See Subsection  2.1.2  Preparation  of  Calibration
Standards.)

5.1.2   Preparation  of Reagent Blank - The reagent blank is prepared
in essentially the same manner  as the field samples.  The difference
in  procedure  occurs  at  the  first step.  In preparing the reagent
blank,  25 ml of absorbing  reagent  is transferred to a 50-ml Class-A
volumetric  flask.   A  25-ml  pipet  may  be used for measuring  and
dispensing the reagent solution; however,  the  use  of  a  graduated
cylinder  will  give  results of acceptable accuracy  and  precision.
After introducing the absorbing reagent  into  the  volumetric flask,
add  water  to the mark, and mix the contents of the flask well.  The
remaining steps for preparing the  reagent  blank  are  identical  to
those of Step 5 under Preparation of Field Samples.

    The reagent blank is used to adjust the analytical results of the
field samples for matrix effects of the  absorbing  reagent  and  the
water.    (The  sample  matrix is simply the medium that contains  the
substance  to  be analyzed, which in this case is nitrate.)   Because
ion chromatography involves separation of the ions prior to detection
and quantification, the potential for the sample matrix  to interfere
with the analysis is small.  For Method 7A,  matrix effects can arise
from the  presence of (a) nitrate contaminant in either the absorbing
reagent  or the water, or (b) a contaminating substance appearing  on
the  chromatogram  at  about the same time as the nitrate  peak.   In
practice, the ion chromatogram should exhibit no significant response
at that point where  nitrate should appear.  Nevertheless, since data
are adjusted  for  the reagent blank, quality results can be obtained
even  if  contamination   exists.   The  presence  of  contamination,
however, indicates the need for greater quality control in connection
with reagent integrity.

5.1.3   Quality  Assurance  Audit Samples - The quality of analytical
results  can  be assessed by  analyzing  nitrate  standard  solutions

-------
                                                   Section No. 3.14.5
                                                   Date July 1, 1986
                                                   Page 4

prepared  by an independent laboratory.  For such standard solutions,
or quality assurance  audit  samples, the concentrations are known to
the control agency (the auditor) but are unknown to the analyst.

    Subsection 3.3.5 of the Federal Register  promulgation  of Method
7A (see Section 3.14.10) requires the analysis of  quality  assurance
audit samples  as described in Method 7.  This means that when Method
7A is used to demonstrate  compliance  with an EPA pollutant emission
standard (specified in 40 CFR Part 60),  a  performance audit must be
conducted  on the analytical phase of the method.  Nitrate samples in
glass vials must be obtained  for  this  performance  audit  from the
Quality  Assurance  Management Office at each EPA Regional Office  or
from  the  responsible  enforcement agency.  The addresses of the EPA
Regional Quality Assurance Coordinators  are  shown  in  Table 5.1 of
Section 3.0.5 of this Handbook.

    The  concentration of each audit sample measured by  the  analyst
must  agree  within  10 percent (relative error) of the actual  audit
concentration.   The relative error is calculated using the following
equation:
                                                         Equation 5-1

                   RE = Cd " Ca x 100

                          Ca

where

         C, = Determined audit sample concentration, mg/dscm, and

         C  = Actual audit sample concentration, mg/dscm.


5.1.4   Analysis  of  Calibration  Standards,  Reagent  Blank,  Field
Samples,  and  Quality  Assurance Samples - Field samples  should  be
recovered  within  4  days  of  sample  collection.     As  currently
written, the method states that the samples should  be stored no more
than 4g days between  collection and  analysis.   However,   a recent
study   utilizing samples from nitric acid  plants  and  power plants
indicates that the storage period between recovery and collection may
be  extended  to 30 days.  Sample analysis using an ion chromatograph
is a straightforward  operation provided that the instrument has been
properly  set  up  (see Section 3.14.2).   All  samples  (calibration
standards,  reagent  blank,  field  samples,  and  quality  assurance
samples)  should be introduced into the ion chromatograph  using  the
same procedure.   Sample  introduction  involves  filling  a constant
volume  sample  loop  using  a syringe or automatic sampling  device.
Sample  loops  give extremely repeatable injection volumes;  however,
the volumes that identify sample  loop  capacity  are not necessarily
accurate.  Nevertheless, accurate results  can  be  obtained  without
having  accurately  known sample loop volumes, provided that the same
sample  loop  is  used  for injecting field samples  and  calibration
standards.   With  this  procedure,  any  inaccuracy in the injection
volume is accounted for by the calibration.

-------
                                                   Section No. 3.14.5
                                                   Date July 1, 1986
                                                   Page 5

    Ion chromatographic  analysis  of  calibration  standards,  field
samples,  reagent blank, and quality assurance samples are  performed
in five phases during the same day,  alternating between the calibra-
tion standards and unknown  samples  to  account for instrument cali-
bration  drift.  These phases are shown in the schedule below.   When
Method 7A is used to demonstrate compliance  with  an  EPA  pollutant
emission  standard, the quality assurance audit samples described  in
Subsection 5.1.3 must be analyzed with the field samples.

         Phase                      Activity

          1         First analysis of all calibration standards.

          2         First  analysis  of all  field  samples,  reagent
                    blank,   and   quality   assurance  samples,   if
                    applicable.

          3         Second analysis of all calibration standards.

          4         Second analysis of  field samples, reagent blank,
                    and quality assurance samples, if applicable.

          5         Third analysis of all calibration standards.

    The calibration standards are analyzed  in  triplicate; the field
samples, reagent blank, and quality assurance samples  in   duplicate.
Replication of analyses  increases  the accuracy and precision of the
results.  Each chromatogram  obtained  from  the  analysis  should be
documented with the following information:

    •   sample identification,

    •   injection point,
    •   injection volume,

    •   nitrate retention time,
    •   sulfate retention time,

    •   eluent flow rate,

    •   detector sensitivity setting, and

    •   recorder chart  speed.


Figure 5.2 shows an  example  chromatogram having acceptable  documen-
tation.  The injection  volume, eluent flow rate, detector sensitivity
setting, and the recorder chart speed need to be documented only once
for the series of chromatograms if these analytical parameters remain
constant over the course of the Method 7A analysis.

    Retention time is the elapsed time between  when  the   sample  is
introduced into the  ion  chromatograph and when the peak of  interest

-------
                                               Section No. 3.14.5
                                               Date July 1, 1986
                                               Page 6
                                               Field Sample: AP-1
                                            Chart Speed: 1 cm/min
                                            Flow Rate: 1.5 ml/min
                                       Detector: 30 yS full scale
                                                 Injection: 50 yl
                                               N03  3.3 minutes
                    Inject
Figure 5.2.   Example of chromatogram having adequate documentation.

-------
                                                   Section No. 3.14.5
                                                   Date July 1, 1986
                                                   Page 7

occurs.  Peaks on the chromatogram may be qualitatively identified by
retention time.  Retention times  can be easily computed from chroma-
tograms provided that the injection point  is  indicated  clearly and
the chart speed is known.  Identification of the injection  point  is
necessary because a chromatogram's trace will not show when injection
occurred.

    Record the results  for  the  calibration  standards,  the  field
samples, and reagent blank on the appropriate  analytical  data  form
(Figures  5.3 and 5.4, respectively).   As  discussed  in  Subsection
2.1.3 and shown in Figure 5.3, the percent deviation  from  the cali-
bration  curve of the average response  value  for  each  calibration
standard must be calculated and must be within 7 percent.  A detailed
discussion of preparation of the calibration curve and calculation of
the  calibration  factor (S) is found in 3.14.2.  The example data in
Figure 5.3 shows the use of linear regression to calculate  S  and  a
non-zero intercept; the example data in Figure 2.3 shows  calculation
of S with a zero intercept using least  squares.  Equation 2-1 or 2-4
along with Equation  2-2  (repeated  below) are used to calculate the
percent deviation  using  either  a  zero  intercept  (Eq.  2-1) or a
non-zero intercept (Eq. 2-4).

                                                         Equation 2-1
    vig N02   = S (yg/mm) x Detector  (mm)
    Predicted              Response
                              H

                                                         Equation 2-4

    yg N02   = S (ug/mm) /Detector (mm) - I (mm)\
    Predicted            I Response              )
                         x   H

                                                         Equation 2-2

    Deviation  = "9 NO2 Predicted - yg NO2 Standard x 10Q%
       (%)                    vi g N02  Standard


    For the analyses of the field samples,  average  the two  response
values  .of  each sample  (see Figure  5.4).   The  calculated   average
should have units consistent with those of the calibration curve, for
example,  units of peak height, peak area, etc.  The pair of  response
values for each sample must each agree within 5 percent of their mean
for  the  analysis to be valid.  For this computation, the  following
equation is used:

                                                         Equation 5-2

    Deviation  (%) = Instrument Response - Mean Response  x 1QQ%
                              Mean Response

-------
Plant
Date
                retftr  rl**rh
                  I,
Location
Analyst
                                                              Section  No.  3.11.5
                                                              Date  July  1,  1986
                                                              Page  8
Was an integrator used?
                             yes
                                     no
Was the intercept (I) used for calculations? * yes no
Were all points within 7 percent of calculated value? >/ yes
Sample
Identifier
Std 1
Std 2
Std 3
Std 1
Std 5
Sample
Mass
(vg NO,)
25
50
100
150
250
Integrator Response
or Peak Height (mm)
H
1
0.3
/4.3
u,^
30-0
57, 2
              yg NOp/mm
   Predicted Sample Mass (yg N0_)
    yg N02 = H x S = (
                                                           Equation 2-1
Predicted Sample Mass using Linear Regression to Calculate Calibration Factor (S)
  and Non-Zero Intercept (I)
   y = mx + b; m =
                             ;  b =
                                   =  4.30/3 ;
        m
                   m
   y = H; and b = I (Intercept) = 2

  Predicted Sample Mass (yg N02)

   yg N02 = S(H - I)

   yg N02 at 25 yg standard = 4-WI (  &17 - ^.
                                                           Equation 2-4
        Figure 5.3.  Analytical data form for analysis of calibration standards.

-------
                                                          Section No.  3.1^-
                                                          Date July  1,  1986
                                                          Page 9
Date samples received  3//
         i        />
Plant   ,
                                     Date samples analyzed   3
                                             Run number(s) AP~l l 3.
                                                                1 I / & S

                                                               ~l l  3.   STk
Location
                            May} /Wio>  Analyst
  Calibration factor (S)   4-. 3O/     Intercept (I),  if applicable ;%. . ftt

  Reagent blank values:   0.0  1st,  0-0   2nd,  0-0  Avg
Field
Sample
Number
Af-,
4/-Z
Analysis
Number
'£
2-tfp~
Instrument
Response
(mm)
28.7
30.2-
23.7
Mean
Instrument
Response
(mm)
^cl. sr

Deviation of two samples, (%) = 100 x
Percent
Deviation
Z.2-

Mean
Instrument
Response
Blank
Corrected
(H)
V.sr
13.1
Dilution
Factor
(F)
/

Mass of
Field
Sample
//4.4

Al ~ A2 (must be less than 5X)
                              = 100
Mass of field sample
  without intercept
  (US N02)
Mass of field sample
  with intercept

  (Vg N0)
                             = S x H x F
                             = S  (H - I) F
        Figure 5-^«  Analytical data form for analysis of field samples.

-------
                                                   Section No. 3.14.5
                                                   Date July 1, 1986
                                                   Page 10

    The reagent blank is analyzed at the  same time as the field sam-
ples.  The average blank  corrected instrument response (H) is deter-
mined  by  subtracting  the  blank  value from the average instrument
response  for  each sample.  The blank corrected instrument  response
(H),  the  dilution  factor (F), and the calibration factor (S) [with
intercept  (I) if necessary]  are then used  to   calculate the  mass
of N02 per sample as shown in Figure 5.4.

-------
                                                              Section 3-14.5
                                                              Date July 1,  1986
                                                              Page 11
                   Table 5.1.  ACTIVITY MATRIX FOR SAMPLE ANALYSIS
Characteristic
 Acceptance Limits
Frequency and method
    of measurement
   Action if
 requirements
 are not met
Calibration
  standards
Data points for cali-
bration standards must
not deviate from the
linear calibration
curve by more than
+1%
Conduct for all analy-
ses of field samples
and calibration stan-
dards
Remake and reana-
lyze standards for
data points that
do not meet cri-
teria; interpret
data using another
technique  (e.g.
peak area instead
of peak height);
strictly observe
.pipetting tech-
nique; use cali-
bration factor
with y-intercept
for calculations;
calibrate conduc-
tivity detector
Field sample
Results from dupli-
cate analyses must
be within 5 percent
of mean value
Conduct for all
analyses of field
samples
                No results exceeding
                value for calibration
                standard having larg-
                est concentration
                        Applicable to all
                        analyses of field
                        samples; determined
                        by visual inspection
Repeat duplicate
analysis,  and
strictly observe
correct pipetting
technique; seek
assistance with
analytical tech-
nique

Dilute blank and
and affected field
sample with equal
volumes of water
and repeat analy-
ses of both
Performance
 audit of
 analytical
 phase
See Section 3-14.8
See Section 3.14.8
See Section 3.14.8
Data
 recording
All pertinent data
recorded on Figs.  5 • 1,
5.2, 5-3, and 5-4
Visually check
Supply missing
data

-------
                                                   Section No.  3.14.6
                                                   Date July 1, 1986
                                                   Page 1
6.0  CALCULATIONS
    Calculation errors due to procedural or mathematical mistakes can
be a large component of total  system error.  Therefore, it is recom-
mended that each set of  calculations  be  repeated  or spot-checked,
preferably by a team  member  other  than  the  one who performed the
original calculations.  If a difference greater than  typical  round-
off error is detected, the calculations  should  be  checked step-by-
step  until the source of error is found and corrected.   A  computer
program  is  advantageous  in  reducing  calculation  errors.   If  a
standardized computer program is used, the original data entry should
be  checked,  and  if  differences are observed, a new  computer  run
should be made.  Table 6.1 at the end of this section summarizes  the
quality assurance activities for calculations.

    Calculations  should  be  carried  out at least one extra decimal
figure  beyond that of the acquired data, and should be rounded after
final calculation to two significant  digits  for each run or sample.
All  rounding  of  numbers should be performed in accordance with the
ASTM 380-76 procedures.  All calculations are then recorded on a form
such as the one in Figure 6.1A.

6.1  Nomenclature

     The following nomenclature is used in the calculations:

    P_ = final absolute pressure of flask, mm (in.) Hg,

    P. = initial absolute pressure of flask, mm (in.) Hg,

  P .. = standard absolute pressure, 760 mm (29.92 in.) Hg,

    Tf = final absolute temperature of flask, °K (°R),

    T± = initial absolute temperature of flask, °K (°R),

  T .. = standard absolute temperature, 293°K (528°R),

   V   = sample volume at standard conditions, dry basis, ml,
    sc
    Vf =s volume of flask and valve, ml,

    V  = volume of absorbing solution, 25 ml,
     O

     H = sample peak height or area (blank should be subtracted
         out), mm,

     F = dilution factor (required only if additional sample
         dilution was needed to reduce the concentration into
         the range of calibration),

     C = sample concentration of NO  as N00, mg/dscm,
                                   X      ft

-------
                                                   Section No. 3.14.6
                                                   Date July 1, 1986
                                                   Page 2

    S = calibration factor,  y g/mm,  and

    I = intercept term from calibration equation,  mm.

6.2  Calculations

    The following four  Subsections outline the procedures for calcu-
lating the concentration of nitrogen oxides  in  samples.  Subsection
6.2.1 presents the equation for calculating  the  sample  volume on a
dry basis at standard conditions.

    Subsection 6.2.2 presents the equation for calculating the sample
concentration  of  nitrogen oxides as it appears in Method 7A.   This
equation utilizes  the  calibration  factor, S, determined during the
calibration of the ion chromatograph (see Subsection 2.1.3 of Section
3.14.2).  Subsection 6.2.3 offers an alternative approach  acceptable
to  the  Administrator  for  calculating the sample concentration  of
nitrogen  oxides  utilizing the calibration factor, S, and the inter-
cept  term,  I, from the  calibration  equation.   This  equation  is
determined  following  the  procedures  outlined  in  Method  7D  for
calibration of the ion chromatograph (see Subsection 2.1.3 of Section
3.14.2).

    Subsection 6.2.4 presents a simple equation for converting sample
concentration  to  parts  per million.  Examples  of  nitrogen  oxide
calculation forms are presented at the end of each section and should
be used with the appropriate calculation methodology.

6.2.1  Sample Volume - Calculate the sample volume on a dry  basis at
standard conditions  [760  mm  (29.92  in.)  Hg and 293 K (528 R)] by
using the following equation.

               T   (v  - v  \   /P    P v                 Equation 6-1
         V   =  <3t-d* f    a'   '  f    •!
         vsc   —^—	—
std

             P
             = K.(Vf - 25 ml)  /If
                1  f           \T
where
                       °K
         K. = 0.3858 	 for metric units, or
                     mm Hg

         K. = 17.64   °R  for English units.
                   in. Hg

6.2.2  Sample Concentration Using the Calibration Factor, S  - Calcu-
late  the sample concentration on a dry basis at standard  conditions
using the calibration factor, S, as shown in Equation  6-2  when  the
calibration factor S was calculated with  no  intercept.  See Figures

-------
                                                   Section No. 3.14.6
                                                   Date July 1, 1986
                                                   Page 3

6.1A and 6.IB for examples  of  calculation  forms  for  English  and
metric units, respectively.

                                                         Equation 6-2

             HSF x 104
               V
                sc
where
         4
       10  =1:10 dilution times conversion factors of
                 mg          10  ml
                 3              3
               10  y g          m

6.2.3   Sample  Concentration  Using  the  Calibration  Equation  and
Factor, S -  Calculate  the  sample  concentration  on a dry basis at
standard conditions  using  the  calibration factor and the intercept
term for the calibration  equation  as  shown  in  Equation 6-3.  See
Figures 6.1A and 6.IB  for  examples of calculation forms for English
and metric units, respectively.

    r   „ (H-I) SF * 104                                 Equation 6-3

         2~^

where

    K2 = 1 for metric units, or

    K9 = 6.243 x 10~8  dscm/mg for English units.
     ^                 dscf/lb
     4
   10  =1:10 dilution times conversion factors of

                 mg        10  ml
                 3      X     3
               10  yg        m    .

6.2.4  Sample Concentration  in  Parts-Per-Million  - If desired, the
concentration  of  N02  may  be  calculated  as  ppm N02 at  standard
conditions using Equation 6-4 as shown below.

                                                        Equation 6-4
    ppm N02= K3 C

where

    K3 = 0.5228 —ppm N02	 for metric units, or
                 mg NO^/dscm

    Kq = 8.375 x 106 	PPnL_N02	 for English units.
                     Ibs N02/dscf

-------
                                                Section No. 3.14.6
                                                Date July 1, 1986
                                                Page 4
                           Sample Volume


    vf = 2-0'L 3 ml, Pf =  2-7  .  64 in.  Hg,  Tf = £T3 3 °R


     _L   ™"™* *™~   "*~~ ™^          JL    ~~ ~~ ^~

                           p     p  v    ,  -, x.            Equation 6-1
   Vo^ = 17.64 (V, - 25)  {_£ _  ^±_\ = I  7 $ 0 ml
    SC           I.
i)  /If _ M  -'_/7g^
   \ ^p    '^  »
   v if   i±/
                        Sample Concentration

(No Intercept Used)

    H = __ . __ mm, S = ____ yg/mm,


    F ' --- ' vsc ' ---- ml
                                                        Equation 6-2
                             4
    C = 6.243 x 10"8 HSF x 10   =      .      x 10"5 Ibs NO0/dscf
                        V                 ~              z
                         sc


(With Intercept Used)

    H = _2-3 . _/Omm, I = _  Z-'. f5fmm,  S = 4._5'£ / yg/mm,

    F = _ 1-0, Vgc = _/7 ^0 ml

                                                        Equation 6-3
                                  4
    C = 6.243 x 10~8 (H-I)SF x  10  =    3 .  0 f x 1Q-5 lbg N0 /dscf
                           v                                ^
                             sc
                    Sample Concentration in ppm

                                                        Equation 6-4
    ppm N02 = 8.375 x 10  C  = _  2, 5" 5
    Figure 6.1A.  Nitrogen oxide calculation form (English units).

-------
                                                Section  No.  3.14.6

                                                Date July 1,  1986

                                                Page 5
                            Sample Volume




    vf = 2-9. L $ m1'  pf  "  7 0 t. Q mm Hg, Tf =  £ ? 5f. £ °K
    P
± - _ / 6.. 0 mm  Hg,  T± = _/•£ 5_. 5TWK
   Var, = 0.3858  (V,  -  25)   /If _ li | = _/ 7 fc> 5_. ml

    SC            I         I m    m
                             Tf   T±.
                         Sample Concentration



(No Intercept Used)



      H =  _ _• _ _ WMr    S - _ _ _ _ v g/mm



      F = ___, V    =____ m^



                                                         Equation 6-2
                  4

      C = HSF x 10  = _.  	 x 103 mg N02/dscm

              sc





(With Intercept Used)


        H*y "2.   / f\      T       7  Q £~    r*    ri *flt /^ /    /
      = (* 2 • / U mm, I  =    £•-•. / j mm, s =  7~. 3 u i  yg/mm,







                                                         Equation 6-3



    C = (H-I)SF x 10   =  ^) . 4 ^ rx 103 mg N09/dscm

           V
            SC






                    Sample Concentration in ppm





                                                         Equation 6-4
    ppm NO-, = 0.5228  C =   ^-D T ppm N00
          «               ""^ *~" ~~ """"       «b








    Figure 6.IB.  Nitrogen oxide calculation form  (metric units).

-------
                                         Section No. 3.14.6
                                         Date July 1, 1986
                                         Page 6
Table 6.1.  ACTIVITY MATRIX FOR CALCULATIONS
Characteristics
Sample volume
calculation
Sample mass
calculation
Sample concen-
tration
Calculation
check
Document and
report re-
sults
Acceptance limits
All data available;
calculations correct
within round-off error
As above
As above
Original and checked
calculations agree
within round-off error
All data available;
calculations correct
within round-off error
Frequency and method
of measurement
For each sample, exam-
ine the data form
As above
As above
For each sample, per-
form independent cal-
culation using data on
Figs. 4.1, 4.2, and
4.3
For each sample, exam-
ine the data form
Action if
requirements
are not met
Complete the
data, or void
the sample
As above
As above
Check and
correct all
data
Complete the
data, or void
the sample

-------
                                                   Section No.  3.14.7
                                                   Date July 1, 1986
                                                   Page 1
7.0  MAINTENANCE
    The normal use  of emission-testing equipment subjects it to cor-
rosive gases, extremes in temperature, vibration, and shock.  Keeping
the equipment in good operating order over an extended period of time
requires knowledge of the equipment and a routine maintenance program
which should be performed quarterly or upon improper  functioning  of
the apparatus.  As for Method 7, it is suggested that the vacuum pump
be disassembled and cleaned yearly.  A summary of the components with
maintenance procedures  is  presented in Table 7.1 at the end of this
section.   These procedures are not required, but are recommended  to
increase the reliability of the equipment.

7.1  Pumps

    Several types of pumps are used in  the  present commercial samp-
ling  trains.   The  two  most  common are the fiber vane  pump  with
in-line oiler and the diaphragm pump.  The fiber vane pump requires a
periodic  check  of  the  oiler jar.  The oil should be  translucent.
During the yearly disassembly or if the fiber vane pump starts to run
erratically, the head should  be removed and the fiber vanes changed.
The  diaphragm  pump  will  show  a  leak  when  the  diaphragm needs
changing.   If the diaphragm.pump runs erratically, it is usually due
to  a bad diaphragm (causing  leakage)  or  to  malfunctions  in  the
valves.   The  valves  should  be  cleaned   annually   by   complete
disassembly of the pump.

7.2  Shipping Containers

    Since the majority of the sampling  train is glassware, the ship-
ping  containers  are  very important for protection and safety.  All
shipping   containers   should  be  inspected  quarterly  for   their
condition,  and  repaired  or  modified  to  assure the safety of the
equipment.

7.3  Ion Chromatograph

    Maintenance activities and schedules  for  ion chromatographs are
make and model  specific.   It  is  therefore  recommended  that  the
analyst  consult  the  operator's manual for instructions relative to
maintenance practices and procedures.

    Guard columns, while not required,  are  recommended for use with
the ion chromatograph in order to extend column lifetime.

-------
                                                             Section No.  3.1^-7
                                                             Date July 1, 1986
                                                             Page 2
                          Table 7.1.  ACTIVITY MATRIX FOR MAINTENANCE
  Apparatus
 Acceptance criteria
Frequency and method
    of measurement
  Action if
requirements
are not met
Fiber vane pump
Oil translucent; pump
leakless and capable
of pulling a vacuum of
less than 75 mm (3
in.) Hg absolute
pressure
Check oiler jar
periodically; remove
head and change fiber
vanes
Replace as
needed
Diaphragm pump
Leak free, valves
functioning properly,
and capable of pulling
a vacuum of < 75 
-------
                                                   Section No.  3.14.8
                                                   Date July 1, 1986
                                                   Page 1
8.0  AUDITING PROCEDURE
    An  audit is an independent assessment of data quality.  Indepen-
dence is achieved if the individual(s) performing the audit and their
standards and equipment are different from the regular field team and
their standards and equipment.  Routine quality assurance checks by a
field team are necessary  to generate good quality data, but they are
not part of the auditing  procedure.   Table  8.1  at the end of this
section summarizes the quality assurance functions for auditing.
                                               19 20 21
    Based on the results of collaborative tests  '  '    of Method 7,
two specific performance audits are recommended:

    1.  Audit of the analytical phase of Method 7A.
    2.  Audit of data processing.

It is suggested that a systems audit be conducted as specified by the
quality assurance  coordinator,  in  addition  to  these  performance
audits.   The  two  performance  audits  and  the  systems  audit are
described in detail in Subsections 8.1 and 8.2, respectively.

8.1  Performance Audits

    Performance  audits are made to evaluate quantitatively the qual-
ity  of data produced by the total measurement system (sample collec-
tion, sample analysis, and data processing).  It is  recommended that
these audits be performed  by  the  responsible  control  agency once
during every enforcement source test.   A source test for enforcement
comprises a series of runs  at  one source.  The performance audit of
the analytical  phase  is  subdivided  into  two steps: (1) a pretest
audit which is optional,  and  (2) an audit during the field sampling
and/or analysis phase which  is required.  No audit is recommended at
this time for the sample collection phase.

8.1.1  Pretest  Audit  of  Analytical  Phase (Optional) - The pretest
audit described in this section can  be  used  to  determine the pro-
ficiency of the analyst,  the quality of the  standard  solutions  in
the Method  7A  analysis, and the ability to perform the computations
correctly.  It  should  be  performed at the discretion of the agency
auditor, the laboratory  supervisor,  source test company, or quality
assurance officer.  The analytical phase  of Method 7A can be audited
with the use of aqueous potassium or sodium nitrate samples.  Aqueous
sodium nitrate samples  may  be  prepared  using  the  same procedure
described in Section 3.14.2 for calibration standard preparation.

    The  pretest  audit  provides  the  opportunity  for  the testing
laboratory to check the accuracy of  its  analytical procedure.  This
audit  is especially recommended for a laboratory with little  or  no
experience  with  the  Method 7A analysis procedure described in this
Handbook.

    As  an  alternative to preparing their own audit  samples  for  a
pretest audit, a testing laboratory may, 30 days prior to the time of

-------
                                                   Section No. 3.14.8-
                                                   Date July 1, 1986
                                                   Page 2

the planned pretest  audit,  make  a  request  to EPA's Environmental
Monitoring Systems  Laboratory,  Quality  Assurance  Division, Source
Branch, Mail Drop 77A, Research  Triangle Park,  North Carolina  27711
for  known  quality  control  samples.   These  samples  are  aqueous
potassium nitrate samples (and not sodium nitrate samples).

    The relative error for  each  of  two samples should be within 10
percent of true  value.   The relative error (RE) is an indication of
the bias that may be associated with the  analytical  phase of Method
7A.  Calculate RE using Equation 8-1.
                RE = Cd " Ca x 100

                       Ca
                                                         Equation 8-1
where

    C, = Determined audit sample concentration, mg/dscm, and

    C  = Actual audit sample concentration, mg/dscm.
     O

8.1.2  Audit of Analytical Phase of the Field Test  (Required)  -  As
stated in Sections  3.3.9  and 4.4 of 40 CFR 60, Appendix A, Method 7
(49  FR  26522, 06/27/84), when the method is  used  for  enforcement
testing, the analyst must analyze two audit  samples  along  with the
field  samples.   The  testing laboratory should notify  the  respon-
sible agency requiring  the performance test of the intent to test at
least 30 days prior to the  enforcement source test.  The responsible
agency will provide two audit samples to be  analyzed  along with the
field  samples from the enforcement source test.  The purpose of this
audit is to assess the data quality at the time of the analysis.   If
EPA  is  the agency  requiring  the  performance  test,  the  testing
laboratory should notify the Quality Assurance Management  Office  in
the  respective  EPA  Regional  Office.   The addresses  of  the  EPA
Regional Quality Assurance Coordinators  are  shown  in  Table 5.1 of
Section 3.0.5 of this Handbook.

     The two audit samples and the compliance samples must be concur-
rently analyzed in the same manner  to  evaluate the technique of the
analyst, the standards  preparation,  and computation skills.  (Note:
It is recommended that known  quality  control  samples  be  analyzed
prior to the compliance  and  audit  sample  analysis to indicate any
problems.  One source of these samples is the Source Branch listed in
Subsection  8.1.1.)   The  same  analyst,  analytical  reagents,  and
analytical system  shall  be used both for compliance samples and the
EPA audit samples; if this  condition  is met, auditing of subsequent
compliance  analyses  for  the same enforcement agency within 30 days
may not be required.  An audit sample set may not be used to validate
different  sets  of  compliance samples  under  the  jurisdiction  of
different  enforcement agencies, unless prior arrangements  are  made
with both enforcement agencies.

-------
                                                   Section No. 3.14.8
                                                   Date July 1, 1986
                                                   Page 3

     Calculate the concentrations  in  mg/dscm  using  the  specified
sample  volume in the  audit  instructions.   (Note:   Indication  of
acceptable results may be obtained immediately by reporting the audit
results in mg/dscm and compliance  results  in total mg N02/sample by
telephone  to  the  responsible  enforcement  agency.)   Include  the
results of both audit samples, their  identification numbers, and the
analyst's  name  with  the  results  of  the compliance determination
samples  in  appropriate  reports to the EPA Regional Office  or  the
appropriate  enforcement  agency.   Include  this   information  with
subsequent compliance analyses for the same enforcement agency during
the 30-day period.

     The concentration of each audit sample measured by  the  analyst
shall  agree  within  10 percent of the actual concentration.  If the
10-percent specification is not met, reanalyze the compliance samples
and audit samples, and  include  initial and reanalysis values in the
test report.

     Failure to meet the 10-percent specification may require retests
until the audit problems are resolved.  However, if the audit results
do not affect the compliance or noncompliance  status of the affected
facility, the Administrator  may  waive  the  reanalysis requirement,
further audits, or retests and accept  the  results of the compliance
test.   While  steps  are being taken to resolve audit analysis prob-
lems, the Administrator  may also choose to use the data to determine
the compliance or noncompliance  status  of  the  affected  facility.
Other  applications  of  Method  7A (i.e., Performance  Specification
Tests) should follow agency recommended or required procedures.

8.1.3   Audit..Qf2Rai;a Processing - Calculation  errors  are prevalent
in  Method 7.  '  '    Data processing errors can  be  determined  by
auditing  the recorded data on the field and laboratory  forms.   The
original and audit (check) calculations should agree within round-off
error; if not, all of the remaining data should be checked.  The data
processing  may  also be audited by providing the testing  laboratory
with specific  data  sets (exactly as would appear in the field), and
by requesting that the data calculation  be  completed  and  that the
results be returned to the agency/organization.  This audit is useful
in  checking  both computer  programs  and  manual  methods  of  data
processing.

8.2  Systems Audit

     A systems audit is an on-site qualitative  inspection and review
of the total measurement system (sample collection,  sample analysis,
etc.).  Initially, a systems audit  is  recommended for each enforce-
ment source  test,  defined  here  as  a  series of three runs at one
source.  After the test team gains experience  with  the  method, the
frequency  of audit may be reduced--for example, to  once  for  every
four tests.

     The  auditor  should  have extensive  background  experience  in
source  sampling, specifically  with  the  measurement  system  being
audited.  The functions of the auditor are summarized below:

-------
                                                   Section No. 3.14.8
                                                   Date July 1, 1986
                                                   Page 4

     1.    Inform  the testing team of the results of pretest audits,
specifying any area(s) that need special attention or improvement.

     2.   Observe procedures and techniques  of the field team during
sample collection.

     3.   Check/verify records  of  apparatus  calibration checks and
quality control used in the laboratory  analysis  of  control samples
from previous source tests, where applicable.

     4.    Record  the results of the audit, and  forward  them  with
comments to the team management so that appropriate corrective action
may be initiated.

     While  on  site, the auditor observes  the  source  test  team's
overall performance, including the following specific operations:

     1.   Setting up and leak testing the sampling train.

     2.   Preparing the absorbing solution (if performed on-site) and
adding it to the collection flasks.

     3.   Collecting the sample.

     4.    Sample   absorption   procedures,   sample  recovery,  and
preparation of samples for shipment.

Figure 8.1 is a suggested checklist for the auditor.

-------
                                                              Section No. 3.14.8
                                                              Date July 1, 1986
                                                              Page 5
Yes
No
Comment
 /
  /
                         Presampling preparation

                1.    Plant operation parameters variation

                2.    Calibration of the flask and valve volume—three
                      determinations

                3.    Absorbing reagent preparation
                5.
                                On-site measurements

                            Leak testing of sampling train

                            Preparation and introduction of absorbing solution
                            into sampling flask
                                    Postsampling
                              (Analysis and Calculation)

                      6.    Control sample analysis

                      7.    Sample aliquotting techniques

                      8.    Ion chromatographic technique

                              a. Preparation of standard nitrate samples
                                (pipetting)
                              b. Calibration factor (+_ 7 percent for all
                                standards)
                              c. Duplicate sample values within 5 percent
                                of their mean
                              d. Adequate peak separation

                       9. Audit results (+ 10%)

                            a. Use of computer program
                            b. Independent check of calculations
                                      Comments
              Figure 8.1.  Method 7A checklist to be used by auditors.

-------
                                                              Section No. 3.1U.8
                                                              Date July 1, 1986
                                                              Page 6
               Table 8.1.  ACTIVITY MATRIX FOR AUDITING PROCEDURE
Audit
 Acceptance Limits
Frequency and method
    of measurement
  Action if
requirements
are not met
Performance
  audit of
  analytical
  phase
Measured RE of the
audit samples shall
be within 10% for
both audit results
Frequency; Once during
every enforcement source
test*
Method; Measure QA sam-
ples and report values
to responsible agency
Review operating
technique and/or
calibration check
Data
  processing
  errors
Original and checked
calculations agree
within round-off
                error
Frequency;  Once during
every enforcement
source test
Method; Independent
calculations starting
with recorded data on
Figures 4.1 and 5.1
Check and correct
all data for the
audit period rep-
resented by the
sampled data
Systems
  audit—
  observance
  of tech-
  nique
Operational tech-
nique as described
in this section of
the Handbook
Frequency; Once during
every enforcement source
test until experience
gained, then every
fourth test
Method; Observation of
techniques assisted by
audit checklist,
Fig. 8.1
Explain to team
their deviations
from recommended
techniques, and
note on Fig. 8.1
*As defined here, a source test for enforcement of the NSPS comprises a series
 of runs at one source.  Source tests for purposes other than enforcement (e.g.,
 a research project) may be audited at a lower frequency.

-------
                                                   Section No. 3.14.9
                                                   Date July 1, 1986
                                                   Page 1

9.0  RECOMMENDED STANDARDS FOR ESTABLISHING TRACEABILITY

    To achieve data of desired quality,  two essential considerations
are necessary: (1) the measurement  process  must  be  in  a state of
statistical  control  at the time of the  measurement,  and  (2)  the
systematic errors, when combined with the random variation (errors or
measurement),  must result in an acceptable uncertainty.  As evidence
in support of good quality data,  it  is necessary to perform quality
control  checks and independent audits of the measurement process; to
document these data;  and to use materials, instruments, and measure-
ment procedures that can be traced  to  an  appropriate  standard  of
reference.

    Data must be routinely obtained by  repeat  measurements of stan-
dard reference samples (primary, secondary, and/or working standards)
and the establishment of a condition of process control.  The working
calibration  standards  should  be  traceable  to standards of higher
accuracy.

    Class-S weights (made to NBS specifications)  are recommended for
the analytical balance calibration.  See Section 3.6.2 for details on
balance calibration checks.

    Class-A volumetric flasks and pipets (made to NBS specifications)
should be used in the preparation and transfer of solutions.

    Audit samples (as discussed in Section  3.14.8)  must  be used to
validate test results for compliance determination  purposes  and are
recommended  as  an independent check on the measurement process when
the method is performed for other purposes.

-------
  10.0    REFERENCE  METHOD*
                                                                              Section  No.  3.14.10
                                                                              Date  July  1,  1986
                                                                              Page  1
 METHOD 7A—DETERMINATION or  NITROGEN
  OXIDE  EMISSIONS   FROM   STATIONARY
  SOURCES—ION CHROMATOCRAPHIC METHOD
  1. Applicability and Principle.
  1.1  Applicability. This method applies to
 the measurement of nitrogen oxides emitted
 from stationary sources: it may be used as
 an  alternative to Method 7 (as defined in 40
 CFR  Part 60.8(b» to determine compliance
 if the stack concentration is within the ana-
 lytical range. The analytical range of the
 method is from 125  to 1.2SO mg NO./m* as
 NOi (65 to 655 ppm). and higher concentra-
 tions may  be  analyzed  by  diluting the
 sample. The  lower  detection limit  is ap-
 proximately  19 mg/m'  (10 ppm), but may
 vary among Instruments.
  1.2  Principle. A grab sample is collected
 in  an evacuated flask containing  a  diluted
'sulfuric acid-hydrogen  peroxide absorbing
 solution. The  nitrogen  oxides, except ni-
 trous oxide,  are oxidized to nitrate  and
 measured by  ion chromatography.
  2. .Apparatus.
  2.1  Sampling. Same as in Method 7, Sec-
 tion 2.1.
  2.2  Sampling  Recovery.   Same  as  in
 Method 7. Section  2.2.  except the stirring
 rod and pH paper are not needed.
  2.3  Analysis. For the analysis, the follow-
 ing equipment is needed. Alternative instru-
 mentation  and procedures will be allowed
 provided the  calibration precision In Section
 5.2 and acceptable audit accuracy  can be
 met.
  2.3.1  Volumetric Plpets. Class A; 1-.  2-. 4-.
 5-ml  (two for the set of standards and one
 per sample).  6-. 10-. and graduated  5-ml
 sizes.
  2.3.2  Volumetric  Flasks. 50-ml  (two per
 sample  and one per standard), 200-ml, and
 1-llter sizes.
  2.3.3  Analytical  Balance. To measure to
 within 0.1 mg.
  2.3.4  Ion Chromatograph.  The ion  chro-
 matograph should have at least the follow-
 ing components:
  2.3.4.1  Columns.  An anion separation or
 other column capable of resolving  the ni-
 trate ion from sulfate  and  other  species
 present and  a standard  anion suppressor
 column (optional). Suppressor columns are
 produced as proprietary items: however, one
 can be produced in the laboratory using the
 resin available from BioRad Company, 32nd
 and Griffin Streets, Richmond. California.
  2.3.4.2 Pump. Capable  of maintaining a
 steady flow as required by the system.
  2.3.4.3  Flow Gauges.  Capable of measur-
 ing the specified system flow rate.
  2.3.4.4  Conductivity Detector.
  2.3.4.5  Recorder.   Compatible  with  the
 output voltage range of the detector.
   3. ReagenU.
   Unless otherwise indicated. It is Intended
 that  all reagents conform to the specifica-
 tions established by the Committee on Ana-
 lytical Reagents of the American Chemical
 Society, where such specifications are avail-
 able: otherwise, use the best available grade.
   3.1 Sampling. An absorbing solution con-
 sisting of sulfuric acid (H.SO.) and hydro-
 gen peroxide (HiOi)  is required for sam-
 pling. To prepare the  absorbing solution, i
 cautiously add 2.8 ml concentrated H.SO. to
 a 100-ml  flask containing water (same as
 Section  3.2),  and dilute  to  volume with
mixing.  Add 10 ml of  this solution, along
with 6 ml of 3 percent H.O. that has been
freshly prepared from 30 percent solution.
to a  1-llter  flask.  Dilute to  volume with
water and mix well. This absorbing solution
should be used within 1 week of Its prepara-
tion. Do not expose  to extreme heat or
direct sunlight.
  3.2  Sample Recovery. Deionized distilled
water that conforms to American  Society
for Testing and  Materials specification  D
1193-74, Type  3.  Is required for sample re-
covery. At the option  of the analyst, the
KMnO.  test for  oxidizable organic matter
may be  omitted when high concentrations
of organic matter are  not expected  to be
present.
  3.3  Analysis. For the analysis, the follow-
ing reagents are required:
  3.3.1 Water. Same as In Section 3.2.
  3.3.2 Stock Standard Solution. 1 mg NO,/
ml. Dry an adequate  amount of sodium ni-
trate  (NaNOi) at  105  to 110'C for a mini-
mum  of 2 hours just  before preparing the
standard  solution. Then dissolve  exactly
1.847  g of dried NaNO, In water, and dilute
to 1 liter  In  a volumetric flask. Mix well.
This  solution  Is stable for  1 month and
st.ould not be used beyond this time.
  3.3.3 Working  Standard Solution. 25 \t%l
ml. Dilute 5 ml of the standard solution  to
200 ml with water in a volumetric flask, and
mix well. •

  3.3.4 Eluent Solution. Weight 1.018 g  of
sodium carbonate (Na.CO,> and 1.008 g  of
sodium bicarbonate (NaHCO,). and dissolve
in 4 liters of water. This solution is 0.0024 M
Na.COj/0.003 M NaHCO, Other eluents ap-
propriate to the column type and capable of
resolving nitrate  ion from sulfate and other
species present may be used.
  3.35  Quality Assurance Audit  Samples.
Same as required in Method 7.
  4. Procedure.
  4.1  Sampling.  Same as in Method 7, Sec-
tion 4.1.
  4.2  Sample.  Recovery.  Same   as   in
Method 7, Section 4.2. except delete the
steps on adjusting and checking the pH  of
the sample. Do not store the samples more
than 4 days between collection and analysis.
  *  Federal  Register,  Volume  48,  No.  237,  December  8,   1983.

-------
                                                                               Section  No.  3.14.10
                                                                               Date  July  1,   1986
                                                                               Page  2
  4.3  Sample. Preparation. Note the level
of the liquid in the container and confirm
whether any sample was lost during ship-
ment: note this on the analytical data sheet.
If a noticeable amount of leakage  has oc-
curred, either void the sample or use meth-
ods, subject to the approval of the Adminis-
trator, to correct the final results. Immedi-
ately before analysis, transfer the contents
of the shipping container  to a 50-mJ volu-
metric flask, and rinse the container twice
with 5-ml portions of water. Add the rinse
water to the flask, and dilute to the  mark
with water. Mix thoroughly.
  Plpet a 5-ml aliquot of the sample into a
50-ml volumetric flask, and dilute to the
mark with water. Mix thoroughly. For each
set of determinations,  prepare a reagent
blank by diluting 5 ml of absorbing solution
to 50 ml with water. (Alternatively, eluent
solution  may be used in all sample, stand-
ard, and blank dilutions.)
  4.4  Analysis. Prepare a standard calibra-
tion curve according to Section 5.2. Analyze
the set of standards followed by the set of
samples using the same injection volume for
both  standards  and samples.  Repeat this
analysis sequence followed by a final analy-
sis of the standard set. Average the results.

The two sample values must agree within 5
percent of their mean for the anlaysis to be
valid. Perform  this duplicate analysis  se-
quence on the same day. Dilute any sample
and the blank with equal volumes of water
if  the concentration exceeds  that  of the
highest standard.
  Document each sample chromatogram by
listing the following analytical parameters:
Injection  point,  injection  volume,  nitrate
and sulfate retention times, flow rate, detec-
tor sensitivity setting, and recorder chart
speed.
  4.5  Audit Analysis. Same as required in
Method 7.
  5. Calibration.
  5.1  Flask Volume. Same  aa in Method 7,
Section 5.1.
  5.2  Standard  Calibration Curve. Prepare
a series of five standards by adding 1.0. 2.0,
4.0. 6.0. and 10.0 ml of working standard so-
lution (25 fig/ml) to a series  of five 50-ml
volumetric flasks. (The standard masses will
equal 25, 50. 100. 150. and 250  »ig.) Dilute
each  flask to volume with water,  and mix
well. Analyze with the samples as described
in Section 4.4 and subtract the  blank from
each value.  Prepare or calculate a linear re-
gression  plot to the standard  masses in pg
(x-axis> versus their peak height responses
in millimeters (y-axis).  (Take peak  height
measurements with symmetrical peaks:  in
all other cases, calculate peak areas.) From
this curve, or equation, determine the slope,
and calculate its reciprocal to denote as the
calibration factor, S. If any point deviates
from the line by more than 7 percent of the
concentration at that point, remake and re-
analyze that standard. This deviation can be
determined by multiplying S times the peak
height response for each standard.  The re-
sultant concentrations must not differ by
more than 7  percent  from  each  known
standard mass (Le., 25, 50. 100. 150. and 250
tig).
   5.3 Conductivity Detector.  Calibrate ac-
  cording  to  manufacturer's   specifications
  prior to initial use.
   5.4 Barometer. Calibrate against a mer-
  cury barometer.
   5.5 Temperature  Gauge. Calibrate  dial
  thermometers   against   mercury-in-glass
  thermometers.
   5.6 Vacuum Gauge. Calibrate mechanical
  gauges, if used, against a mercury manome-
  ter such as that specified in Section 2.1.6 of
  Method 7.

  5.7  Analytical Balance. Calibrate against
standard weights.
  6. Calculations
  Carry out the  calculations,  retaining at
least one extra decimal figure beyond that
of the  acquired  data. Round  off  figures
after final calculations.
  6.1  Sample Volume. Calculate the sample
volume V. (in ml) on a dry basis, corrected
to standard conditions, using Equation 7-2
of Method 7.
  6.2  Sample Concentration of NO. as NO,.
Calculate  the  sample concentration C (in
mg/dscm) as follows:
           HSP x 10«
     C=  	       Eq. 7A-1
               "•c


Where:
H = Sample peak height, mm
S = Calibration factor. »xg/inm

F = Dilution factor (required only if sample
   dilution was needed to  reduce the con-
   centration into the range of calibration)
104 = 1:10 dilution times conversion  factor
   of
             mg

            10'us
                      10* ml
  If desired, the concentration of NOt may
be calculated as ppm NOi at standard condi-
tions as follows:
      ppm NOj - 0.5228 C     Eq. 7A-2


Where:
0.5228 - ml/mg NO,.

-------
                                                                       Section  No.   3.14.10
                                                                       Date  July  1,  1986
                                                                       Page  3
  7. Bibliography.
  1. Mulik. J. D. and E. Sawicki. Ion Chro-
matographic Analysis of Environmental Pol-
lutants. Ann Arbor. Ann Arbor Science Pub-
lishers. Inc. Vol. 2. 1979.
  2. Sawicki, E.. J. D. Mulik. and E. Wittgen-
stein. Ion Chromatographic Analysis of En-
vironmental  Pollutants.  Ann Arbor, Ann
Arbor Science Publishers. Inc. Vol. 1.1978.
  3. Slemer.  D. D. Separation of Chloride
and Bromide from Complex Matrices Prior
to Ion Chromatographic Determination. An-
alytical  Chemistry 52(12:1874-1877). Octo-
ber 1980.
  4. Small.  H.. T. S. Stevens, and W. C.
Bauman. Novel Ion Exchange Chromatogra-
phic Method  Using Conductimetric Deter-
mination. Analytical Chemistry. 47(11:1801).
1975.
  S. Yu. King K. and Peter R. Westlln. Eval-
uation of Reference Method 7 Flask Reac-
tion Time. Source  Evaluation Society News-
letter. 4(4). November 1979. 10 p.

-------
                                                  Section No. 3.14.11
                                                  Date July 1, 1986
                                                  Page 1
11.0  REFERENCES
      1.  Federal  Register,  Volume  48, No. 237, December 8,  1983.
          Method 7A - Determination  of Nitrogen Oxide Emissions From
          Stationary Sources.

      2.  Small, H. T., S. Stevens,  and  W.  C.  Bauman.   Novel Ion
          Exchange  Chromatographic   Method   Using   Conductimetric
          Determination.  Analytical Chemistry,  47(11):801, 1975.

      3.  Johnson,   E.   L.   and   R.   Stevenson.    Basic  Liquid
          Chromatography.  Varian Associates, Inc., 1978.

      4.  Yost,  R.  W.,  L.  S.  Ettre,  and R. D. Conlon, Practical
          Liquid  Chromatography,  An   Introduction.   Perkin-Elmer,
          1980.

      5.  Smith,  F.  C.,  Jr., and R. C. Chang.  The Practice of Ion
          Chromatography.   John  Wiley  and  Sons,  Inc.,  New York,
          1983.

      6.  Stevens, T. S. and M. A. Langhorst. Agglomerated Pellicular
          Anion-Exchange  Columns for Ion Chromatography.  Analytical
          Chemistry, 54 (6):950, 1982.

      7.  Stevens, T. S., G. L. Jewett, and  R.  A. Bredeweg.  Packed
          hollow    fiber   suppressors   for   ion   Chromatography.
          Analytical Chemistry, 54 (7):1206, 1982.

      8.  Mulik,   J.   D.,  and  E.  Sawicki.   Ion  Chromatography.
          Environmental Science and Technology, 13 (7):804, 1979.

      9.  Stevens, T. S., J. C. Davis, and H. Small. Hollow Fiber Ion
          Exchange  Suppressor  for  Ion  Chromatography.  Analytical
          Chemistry, 53 (9):1488, 1981.

     10.  Stevens,  T.  S.   Packed  fibers  and  new  columns speed,
          simplify   ion  Chromatography.    Industrial  Research  and
          Development, September 1983.

     11.  Gjerde,  D.  T.,  J.  S.  Fritz,   and G. Schmuckler.  Anion
          Chromatography  with Low-Conductivity Eluents.  Journal  of
          Chromatography, 186  (509), 1979.

     12.  Jupille, T., D. Surge,  and  D. Togami.  Ion Chromatography
          uses only one column  to  get  all  the ions.   Research and
          Development 26  (3):135, 1984.

     13.  Jenke,   D.    Anion  Peak  Migration  Ion  Chromatography.
          Analytical Chemistry, 53 (9):1535, 1981.

-------
                                             Section No. 3.14.11
                                             Date July 1, 1986
                                             Page 2

14.  Skoog, D. A., and D. W. West.  Fundamentals  of  Analytical
     Chemistry,  Second  Edition.   Holt,  Rinehart and Winston,
     Inc., New York, 1969.

15.  Yu, King  D. and Peter R. Westlin.  Evaluation of Reference
     Method 7 Flask Reaction Time.   Source  Evaluation  Society
     Newsletter, 4(4),  November  1979.   10 p. (Sees. Ill, 114,
     and 301(a)  of  the  Clean Air Act, as amended (42 U. S. C.
     7411, 7414, and 7601(a))).

16.  Steinsberger,  S.  C.  (Entropy  Environmentalists,  Inc.).
     Unpublished results of NO   sample  stability  study.  June
     1987.                    x

17.  Siemer,   D.  D.  Separation of Chloride  and  Bromide  from
     Complex  Matrices  Prior  to Ion Chromatographic Determina-
     tion.  Analytical Chemistry.  52  (12):1874-1877,   October
     1980.

18.  Eubanks,  D.  R.,  and   J.   R.   Stillian.  Care  of  Ion
     Chromatography  Columns.  Liquid Chromatography.  2 (2):74,
     1984.

19.  Hamil,  Henry  F. et. al.  The Collaborative Study  of  EPA
     Methods 5, 6, and 7 in Fossil Fuel  Fired Steam Generators.
     Final Report, EPA-650/4-74-013, May 1974.

20.  Hamil,  H. F., and R. E. Thomas.   Collaborative  Study  of
     Method  for the Determination of Nitrogen  Oxide  Emissions
     from    Stationary   Sources    (Nitric    Acid    Plants).
     EPA-650/4074-028, May 1974.

21.  Hamil,  Henry F.  Laboratory and Field Evaluations  of  EPA
     Methods 2, 6,  and  7.   Final  Report,  EPA  Contract  No.
     68-02-0626,   Southwest  Research Institute,  San  Antonio,
     Texas, October 1973.

-------
                                                  Section No.  3.14.12
                                                  Date July 1,  1986
                                                  Page 1
12.0  DATA FORMS
    Blank data forms  are  provided  on  the  following pages for the
convenience of the  Handbook user.  Each blank form has the customary
descriptive title centered at the top  of  the  page.   However,  the
section-page  documentation in the top right-hand corner of each page
has  been  replaced with a number in the lower right-hand corner that
will  enable  the  user to identify and refer to a similar  filled-in
form in a text section.  For example, Form M7A-1.2 indicates that the
form is Figure 1.2 in  Section  3.14.1  of  the  Method  7A  section.
Future  revisions of these forms, if any, can be documented by  1.2A,
1.2B,  etc.   Eleven of the blank forms listed below are included  in
this section.  Four are  in the Method Highlights subsection as shown
by the MH following the form number.


                   Title

                   Procurement Log

                   Analytical Balance Calibration Form

                   Analytical Data Form for Analysis of
                   Calibration Standards

3.1 (MH)           Pretest Sampling Checks

3.2 (MH)           Pretest Preparations

4.1A AND 4.IB      Nitrogen Oxide Field Data Form (English
                   and metric units)

4.2A and 4.2B      NO  Sample Recovery and Integrity Data
                   Form (English and metric units)

4.3 (MH)           On-site Measurements

5.1 (MH)           Posttest Operations

5.4                Analytical Data Form for Analysis of
                   Field Samples

6.1A and 6.IB      Nitrogen Oxide Calculation Form   (English  and
                   metric units)

8.1                Method 7A Checklist to be Used by Auditors

-------
PROCUREMENT LOG
Item description

«ty.

Purchase
order
number

Vendor

Date
Ord.

Rec.

Cost

Disposition

Comments

                   Quality Assurance Handbook M7A-1.2

-------
               ANALYTICAL BALANCE CALIBRATION FORM
Balance name
Number
Classification of standard weights



Date

0.5000 g

1.0000 g

10.000 g

50.0000 g

100.0000 g

Analyst

                                  Quality Assurance Handbook M7A-2.1

-------
      ANALYTICAL  DATA FORM FOR  ANALYSIS OF CALIBRATION STANDARDS
 Plant

 Date
             Location
             Analyst
 Was an integrator used?
yes
        no
Was the intercept (I) used for calculations? yes no
Were all points within 7 percent of calculated value? yes
Sample
Identifier
Std 1
Std 2
Std 3
Std 4
Std 5
Sample
Mass
(u§ NO.,)
25
50
100
150
250
Integrator Response
or Peak Height (mm)
H
1





2





3





Avg





Predicted
Sample Mass
(yg N00)





no
Deviation
(%)





Predicted Sample Mass using Least Squares to Calculate Calibration Factor (S)
  with Zero Intercept
S = S1H1
H 2
Hl '
Sv
—
S =
3redicted
u g NO- =
+ S2H2 + S^H^ + S^ + S^H^
2222
H H2^ + H^ * Hf + H^
)( ) + ( )( ) + ( )( ) + ( )( ) * ( )( )

P P P 5 P
( r + ( r + ( )2 + ( )2 + ( )2
yg N0_/mm
Sample Mass (yg NOp)
H x S = ( ) x ( ) = Equation 2-1
Predicted Sample Mass using Linear Regression to Calculate Calibration Factor  (S)
  and Non-Zero Intercept (I)

    y =  mx + b; m = 	
        m
                  m
    y = H;  and b = I (Intercept)  = 	

   Predicted  Sample Mass (yg N0_)

    yg N02  =  S(H - I)

    yg N02  at 25 yg standard = 	(
                               Equation 2-
                                     Quality Assurance  Handbook M7A-2.3

-------
                        NITROGEN  OXIDE FIELD DATA FORM (ENGLISH UNITS)
Plant
Sample location
Operator 	
City
Date
Barometric pressure (P,  )
in.  Hg
Sample
number

Sample
point
location

Sample
time
24-hr

Probe
temperature ,
OF

Flask
and valve
number

Volume
of flask
and valve {Vp) ,
ml

Initial pressure
in. Hg
Leg A±

Leg Bi

Pia

Initial temperature
°F(ti)

°R(Ti)b

= p
   bar
       460°F.
                                                        Quality Assurance Handbook M7A-4.1A

-------
   Plant
   Sample location



   Operator 	
 NITROGEN OXIDE FIELD DATA  FORM (METRIC UNITS)


	  City 	



                         Date
                         Barometric pressure (P.  )
                                              oar
mm Hg
Sample
number

Sample
point
location

Sample
time
24-hr

Probe
temperature ,
°C

Flask
and valve
number

Volume
of flask
and valve (VF) ,
ml

Initial pressure
in. Hg
Leg A±

Leg B±

*ia

Initial temperature
°C(ti)

°R(Ti)b

pi = pbar
         273°c-
                                                           Quality Assurance  Handbook  M7A-4.1B

-------
         NO   SAMPLE RECOVERY AND INTEGRITY DATA FORM (ENGLISH UNITS)
    Plant                                  Date
    Sample recovery personnel
     Barometric pressure,  (Pbar)
in.  Hg
    Person with direct responsibility  for recovered samples
Sample
number

Final pressure,
in. Hg
Leg Af

Leg Bf

Pfa

Final temperature,
°F (tf)

°R (Tf)b

Sample
recovery
time,
24-h

Liquid
level
marked

Samples
stored
in locked
container

f = pbar - -       ^'f =  fcf + 460°F-
Lab person with direct responsibility for recovered samples
Date recovered  samples received 	  Analyst 	
All samples identifiable? 	
Remarks
All liquids  at marked level?
Signature of lab sample trustee
                                                  Quality Assurance Handbook  M7A-4.2A

-------
        N0x SAMPLE RECOVERY  AND INTEGRITY  DATA  FORM (METRIC  UNITS)
    Plant
                                           Date
    Sample  recovery personnel
          Barometric pressure, (P,   )
                                oar
    Person with direct responsibility for recovered samples
                                                                                mm Hg
Sample
number

Final pressure,
mm Hg
Leg Af

Leg Bf

V

Final temperature,
°c (tf)

°K (Tf)b

Sample
recovery
time,
24-h

Liquid
level
marked

Samples
stored
in locked
container

  = Pbar - (Af
tf + 273°C.
Lab person with direct responsibility for recovered samples



Date recovered samples received 	  Analyst 	



All samples identifiable? 	



Remarks
      All liquids at marked  level?
Signature of lab sample trustee
                                                    Quality Assurance  Handbook M7A-4.2B

-------
        ANALYTICAL DATA FORM FOR  ANALYSIS OF FIELD SAMPLES
  Date samples received

  Plant
       Date samples analyzed
                Run number(s)
  Location
  Calibration  factor  (S)

  Reagent blank values:
1st,
Analyst 	

Intercept (I),  if applicable

   2nd, 	 Avg



Field
Sample
Number





Analysis
Number




Instrument
Response
(mm)



Mean
Instrument
Response
(mm)





Deviation
(yg N02)

Mean
Instrument
Response
Blank
Corrected
(H)




Dilution
Factor
(F)



Mass of
Field
Sample
(yg N02)

Deviation of two samples, (%) =
         A   - A  i
  100 x  1    2\  (must be less than
                                      Al  + A2
                             = 100
Mass of field sample
  without intercept
      N02)
Mass of field sample
  with intercept
  (yg N02)
= S x H x F
= S (H - I)  F
                                    Quality Assurance Handbook M7A-5.4

-------
          NITROGEN OXIDE CALCULATION FORM (ENGLISH UNITS)




                           Sample Volume



    Vf =	ml,  Pf =	.	in.  Hg,  Tf =	°R


    P± =	.	in.  Hg,  T± =	°R



                                                       Equation 6-1
Vsc = 17.64 (V£ -25)   If _ li  =	
                                              ml
                        Sample Concentration


(No Intercept Used)


    H =	.	mm,  S =	vg/mm,



    F	.  Vsc =	ml


                                                       Equation 6-2
                             4

    C = 6.243 x 10"8 HSF x 10  =     .      x 10"5 Ibs N00/dscf
                        V                               ^
                         sc



(With Intercept Used)


    H =	.	mm,  I =	.	mm,  S =	vg/mm,



    F '	'  Vsc =	ml

                                                       Equation 6-3
                                 4

    C = 6.243 x 10"8 (H-DSF x 10  =     ^      x 1Q-5 lbs N0 /dscf

                           V                                z
                            sc
                    Sample Concentration in ppm




    ppm N02 = 8.375 x 10  C =	ppm N02
                     -                              Equation 6-4
                     D
                                 Quality Assurance Handbook M7A-6.1A

-------
           NITROGEN OXIDE CALCULATION FORM (METRIC UNITS)
                           Sample Volume
'f	
                 ml, Pf =	.
                  mm Hg,  T.  = _
i	•  _
   V   = 0.3858 (V- - 25)
    SC            I
                            _ mm Hg,  Tf
                              .    °K
                                         .  ml
                                                       Equation 6-1
                        Sample Concentration
(No Intercept Used)
      H =	.	mm,    S =	yg/mm
      F -	< vsc •	ml
      C =
          HSF x 10
              sc
                  4
                          x 10  mg NO0/dscm
(With Intercept Used)
    H =     .      mm, I
    F =        V   =
    r	'  vsc
    c a (H-I)SF x 10
           Vsc
                                                       Equation 6-2
                   =	.	mm, S =	y g/mm,
                  	ml
                                                  Equation 6-3
                                 3
                   _ .	x 10  mg N02/dscm
                    Sample Concentration in ppm
    ppm N02 = 0.5228 C =	ppm N02
                                                       Equation 6-4
                                 Quality Assurance Handbook M7A-6.1B

-------
                 METHOD 7A  CHECKLIST TO  BE USED BY  AUDITORS
Yes
No
Comment
                     Presampling preparation


                     1.    Plant operation parameters variation

                     2.    Calibration of the flask and valve  volume	three
                           de terminations

                     3.    Absorbing reagent preparation
                               On-site measurements

                     4.    Leak testing of sampling train

                     5.    Preparation and introduction of  absorbing solution
                           into sampling flask
                                   Postsampling
                             (Analysis and Calculation)

                      6.    Control sample analysis

                      7.    Sample aliquotting techniques

                      8.    Ion chromatographic technique

                             a. Preparation of standard  nitrate  samples
                               (pipetting)
                             b. Calibration factor (+_ 7  percent  for  all
                               standards)
                             c. Duplicate sample values  within 5 percent
                               of their mean
                             d. Adequate peak separation

                       9. Audit results (+_ 102)

                           a. Use of computer program
                           b. Independent check of calculations
                                     Comments
                                        Quality  Assurance Handbook M7A-8.1

-------
                                                Section No.  3.15
                                                Date July 1,  1986
                                                Page 1
                         Section 3.15

         METHOD 7D - DETERMINATION OF NITROGEN OXIDE
              EMISSIONS FROM STATIONARY SOURCES

     (Alkaline-Permanganate - Ion Chromatographic Method)
                           OUTLINE


                                                     Number of
       Section                      Documentation      pages

SUMMARY                                  3.15           3

METHOD HIGHLIGHTS                        3.15           8

METHOD DESCRIPTION

   1.  PROCUREMENT OF APPARATUS
       AND SUPPLIES                      3.15.1        18

   2.  CALIBRATION OF APPARATUS          3.15.2        20

   3.  PRESAMPLING OPERATIONS            3.15.3         6

   4.  ON-SITE MEASUREMENTS              3.15.4        10

   5.  POSTSAMPLING OPERATIONS           3.15.5        13

   6.  CALCULATIONS                      3.15.6         5

   7.  MAINTENANCE                       3.15.7         3

   8.  AUDITING PROCEDURES               3.15.8         6

   9.  RECOMMENDED STANDARDS FOR
       ESTABLISHING TRACEABILITY         3.15.9         1

  10.  REFERENCE METHODS                 3.15.10        9

  11.  REFERENCES                        3.15.11        2

  12.  DATA FORMS                        3.15.12       11

-------
                                                     Section  No.  3.15
                                                     Date  July 1,  1986
                                                     Page  2
                               SUMMARY
    For  EPA Method 7D , an integrated, metered  sample   is   extracted
via a heated probe positioned  at  a  point within  the  duct or stack.
The  sample  is  passed  through  a  series  of   3  restricted orifice
impingers each containing  an  absorbing solution of  sodium hydroxide
( NaOH ) and potassium permanganate  ( KMnO. ) .   The  absorbing solution
reacts with nitrogen oxides  in the effluent gas to form nitrate ion,
   ~
N0«,  and  nitrite  ion, N02  •  Nitrogen oxides  (NO  )  are the sum of
nitric oxide (NO)  and  nitrogen dioxide (N02) whichxare  usually at a
ratio of 19 to  1  by  weight,  respectively, in  the  emission stream.
The collected sample  is  required  to  sit   for   36  hours  prior to
analysis in order for  the  N02~  to  react   completely to N03~.  Ion
chromatography   is   then   used  to  quantify   the  N03~  wnich  is
functionally related to the NO  concentration of  the  effluent sample.
                              a

    The absorbing solution also reacts with  carbon dioxide,   C07,   in
the effluent  sample.   Therefore, EPA Method 3 determinations of C02
must  be  conducted  with Method 7D in order to correct the Method 7D
volumetric data for the volume of C02 absorbed.
    Ammonia, NH3, interferes with Method 7D by causing  NO
be  biased high.  Method 7D results can be  corrected   for"-
using data from concurrent determinations of NH,, .
                                                            results to
                                                             the  bias
    Collection of the NO is presumed to  involve  oxidation -  reduction
reactions  where the NO is oxidized sequentially to  NO~  ~and then  to
*T^N     ml_ — > 1_ — ^ .f — . — — — -1_ J — . __ — .^ ~_ — _L.1_. » ^ ,•. _._ ,_. J_ J ... _»  ... *£ »T^X   _. __ v A ™*
NO
       The half reactions for the formation of N0~   ar
    Mn0
          + 2H20
                   3e  =
    3NO + 60H
                 3NO,
                       + 3H20
and the overall reaction is:
    3NO + Mn04  + 20H
                         3N0
                            40H

                            3e~



                            - MnO,
                            * *)    • J**Vx«  » 11 A V

The half reactions for the formation of NO,

                                  6e"

                                  80H~
                                              are:
    3N0
  '2

2MnO,
            60H~ = 3N0
             4H00
               2
                    6e~ = 2Mn0
        J *  •  ~XA« A W  • W W  — **fc-»**^//^


and the overall reaction is:
3N0
            2Mn0
                                   2Mn0
                                            20H
Reaction S-l

Reaction S-2



Reaction S-3



Reaction S-4

Reaction S-5



Reaction S-6
NO
       A  *  4*l~lLl\*/ *   • A A OV"/ — ^/*^V^Q   • *** ** A xx A


    The  sum of Reactions S-3 and S-6 describes the  reaction of NO to
    *
  3 '

    NO + MnO,
            '4  = N°3
                           0
                           2
                                                          Reaction S-7

-------
                                                    Section No. 3.15
                                                    Date July 1, 1986
                                                    Page 3

    The  rate  of  the  reaction of NO to NO-" is controlled  by  the
solubility of NO.  It takes  approximately  36 hours for the reaction
of N02~ to  N03"  to  reach  completion; the factors controlling this
reaction are unknown.
    Absorption  of  N02  is  also  presumed  to  involve an oxidation
-reduction  reaction.   In  contrast  to NO, N02 is rather  reactive;
thus, it is reasonable to show N02 reacting directly  to  N0o~*   Tne
half reactions are:

    3N02 + 60H~ = 3N03~ + 3H20 + 3e~                     Reaction S-8

    Mn04~ + 2H20 + 3e~ = Mn02 + 40H~                     Reaction S-9

and the overall reaction is:

    3NO0 + MnO.," + 2OH~ = 3NO~ + MnO,, + H00            Reaction S-10
       24              322

    The absorption of C02 involves the simple acid-base reaction with
OH~ to form bicarbonate Ion, HC03~:

                   C02 + OH" = HC03~                    Reaction S-ll


In the  strongly basic absorbing solution, the bicarbonate ion reacts
further to carbonate ion, C03~:

            HC03~ + OH" = H20 + C03=                    Reaction S-12

    Method 7D is applicable to the measurement  of  NO   emitted from
sources in the following categories:                  x

    (a) fossil-fuel-fired steam generators subject to 40 CFR
        Part 60, Subpart D;

    (b) electric utility steam generating units subject to 40 CFR
        Part 60, Subpart Da; and

    (c) nitric acid plants subject to 40 CFR Part 60, Subpart G.

    It may be used as an alternative  to  Method  7 [as defined in 40
CFR Part 60.8(b)] to determine compliance if the  stack concentration
is within the  analytical  range.   The  lower limit ofgdetectability
(with  NO  defined as N02) is approximately 13 mg N02/m  (7 ppm  N02)
when  sampling  is conducted at a flow rate of 500 cc/min for 1 hour.
The  method's  upper  analytical  limit  has  not  been  established;
however, results of  field  evaluations  have  shown  that NO  can be
collected quantitatively at concentrations of  1,782  mg  N02^m  (932
ppm  N02) when sampling is conducted at a flow rate of 500 cc/min for
1 hour.

-------
                                                    Section No. 3.15
                                                    Date July 1, 1986
                                                    Page 4

    The method description which follows is based on the method  that
was promulgated on September 27, 1984.

    Section 3.15.10  contains  a  copy  of  Method 7D, and blank data
forms are provided in Section 3.15.12  for  the  convenience  of  the
Handbook user.

-------
                                                    Section No. 3.15
                                                    Date July 1, 1986
                                                    Page 5
                          METHOD HIGHLIGHTS
    Section 3.15 contains  the  required  procedure  for sampling and
analyzing emissions of nitrogen oxides from stationary sources  using
Method  7D.    For  the method, an integrated sample is taken from  a
point in the duct  or  stack  using  a  heated  probe  constructed of
borosilicate glass, stainless steel, or Teflon™.  The effluent sample
stream  is  passed  through  a  series  of  three  restricted orifice
impingers,  each  containing  200  ml  of a 4.0% (w/w) KMn04 and 2.0%
(w/w) NaOH solution, termed  "alkaline  permanganate  solution."  The
alkaline permanganate  solution  quantitatively removes NO , C02, and
SO-  from  the  effluent  sample stream  and converts  (provided  the
samples  are  allowed to sit for at least 36 hours)  these  to  ions:
N03 , C03~, and SO.",  respectively.   Sampling  is  conducted  at  a
measured flow  rate  between  400 and 500 cc/min for 60 minutes.  The
measured  flow  rate is  on  a  moisture-  and  CO.-free  basis,  and
consequently, when the method is applied to effluents from combustion
processes, the measured flow rate will be less than the sampling flow
rate.  In addition, sampling for C02 must be conducted using Method 3
in conjunction with Method 7D in oraer to correct the volumetric data
for the volume of C02 absorbed.

    After  acquisition, the sample is allowed to sit for a minimum of
36 _hours to ensure that the N02  has been quantitatively reacted  to
NO3  .    Sample   preparation   entails  destruction  of  the  excess
permanganate and filtration of the solid, manganese reaction product,
manganese  dioxide  (Mn02).   NO   as  N03~ is quantified  using  ion
chromatography (1C).            x

    Ion chromatography  is  a  relatively  recent analytical develop-
ment.   The  reader  is  referred12to  the  literature  for  detailed
descriptions  of  the  subject.       Small,  et al.,  developed  the
technique  using the principles of ion  exchange  chromatography  and
conductimetric  detection.   Previous  attempts  to  use this type of
detection were unsuccessful because of the presence of the background
electrolyte used for  elution  of  the ionic species.  Small, et al.,
used a novel combination of resins  to  separate the ions of interest
and neutralize the eluent from the background.

    The • aqueous sample is introduced into a fixed-volume sample loop
by using a plastic  syringe.   Once  injected,  the sample is carried
through  a  separation  column  at different rates according to their
relative  affinities  for   the  resin  material  and  are  therefore
separated into discrete bands.   The  separated  ions are then passed
through a post-separation suppressor device which converts the eluent
ions  into  a  less conducting weak acid while converting the analyte
ions into a highly  conducting  form.   This  permits  the  use  of  a
conductivity cell as a very sensitive detector of all ionic species.
                   12
    Gjerde, et al.,   described a modified ion chromatographic method
that eliminates  the  need  for  a  suppressor  device.   Anions  are

-------
                                                    Section No. 3.15
                                                    Date July 1, 1986
                                                    Page 6

separated  on  a column containing an anion-exchange resin with a low
exchange capacity.   Because  of  the  low  capacity,  a  very dilute
solution of an aromatic organic acid salt may be used as  the eluent.
The conductance of the eluent is sufficiently low that no suppression
is needed.

    For  Method  7D, either suppressed or non-suppressed  1C  may  be
used.    The   basic  ion  chromatograph  will  have  the   following
components:

    (a)  sample injection device,

    (b)  anion separation column,

    (c)  anion suppressor column,  either  packed  bed  or fiber type
         (not required for non-suppressed 1C),

    (d)  conductivity detector, and

    (e)  recorder.

    The critical aspects of the method are (a) the measurement of the
gaseous  sample volume, and (b) the preparation  of  the  calibration
standards for the ion chromatograph.  Analysts are advised to observe
specified  procedures  carefully  at  these  points  of  the  method.
Analysts performing the  method  should be well trained in the use of
the ion chromatograph.

    Collaborative  testing  has been performed for Method 7D and  the
results exhibit accuracy and precision similar to that of Method 7.

    The approporiate blank data forms at the end of this section  may
be removed from the Handbook and used in the pretest, on-site, and in
posttest operations.  Each form has  a subtitle to assist the user in
finding a similar filled-in  form  in the method description.  On the
blank and filled-in forms, the  items/  parameters that can cause the
most significant errors are designated with an asterisk.

1.  Procurement of Apparatus and Supplies
    Section  3.15.1  (Procurement  of  Apparatus  and Supplies) gives
specifications,  criteria,  and  design  features  for  the  required
equipment  and  materials.  The sampling apparatus of Method  7D  has
design features similar to those of Method 6.  Section  3.15.1 can be
used as a guide for procurement  and  initial checks of equipment and
supplies.  The activity matrix (Table 1.1) at the  end of the section
is a summary of the details given  in  the  text and can be used as a
quick reference.

2.  Pretest Preparations
    Section 3.15.2 (Calibration of Apparatus) describes  the required
calibration  procedures and considerations for the Method 7D sampling

-------
                                                    Section No.  3.15
                                                    Date July 1,  1986
                                                    Page 7

equipment  (essentially  the  same  as Method  6)  and  for  the  ion
chromatograph  (the  same as for Method 7A).  Required accuracies for
each component are also included.   A  pretest checklist (Figure 2.5,
Section 3.15.2)  or  a  similar  form should be used to summarize the
calibration  and  other  pertinent  pretest  data.   The  calibration
section may be removed along with the corresponding sections from the
other  methods  and  made into a separate quality assurance reference
manual for use by personnel involved in calibration activities.

    Section 3.15.3 (Presampling Operations) provides  the tester with
a guide  for  equipment  and supplies preparation for the field test.
With the  exception of the preparation of certain reagents, these are
the  same  as  for Method 6 and Method 3.  A pretest preparation form
(Figure 3.1, Section 3.15.3) can be used as an equipment checkout and
packing list.  The method  of  packing  and  the use of the described
packing containers should  help protect the equipment, but neither is
required by Method 7D.

    Activity  matrices  for  the calibration  of  equipment  and  the
presampling operations  (Tables  2.1  and  3.1)  summarize the activ-
ities.

3.  On-Site Measurements
    Section  3.15.4  (On-Site  Measurements)   contains  step-by-step
procedures  for  sample  collection  and for sample recovery.  Sample
collection is similar  to  Method  6,  with  the  exception  that the
alkaline  permanganate  solution  is  placed  in  restriced   orifice
impingers and the C02 content of the stack gas  must be determined to
correct the sample volume for the C02 removed by the  sampling train.
The  on-site  measurement  checklist  (Figure  4.4,  Section  3.15.4)
provides  the  tester  with  a  quick  method of checking the on-site
requirements.   Table 4.1 provides an activity matrix for all on-site
activities.

4.  Posttest Operations
    Section3.15.5fPostsampling  Operations)  gives  the  posttest
equipment  procedures  and a step-by-step  analytical  procedure  for
determination of NO , expressed as N02.  The posttest operations form
(Figure  5.4,  Section  3.15.5)  provides  some  key parameters to be
checked by the tester  and  laboratory  personnel.   The step-by-step
analytical procedure description  can  be  removed  and  made  into a
separate  quality  assurance  analytical  reference  manual  for  the
laboratory personnel.  Analysis of a control sample is required prior
to the analysis of the field samples.  This analysis of independently
prepared, known standards will  provide  the  laboratory with quality
control  checks  on  the  accuracy  and precision of  the  analytical
techniques.   Strict adherence to the Method 7D analytical procedures
must be observed.

    Section  3.15.6  (Calculations)  provides  the  tester  with  the
required equations, nomenclature,  and  significant  digits.   It  is

-------
                                                    Section No. 3.15
                                                    Date July 1, 1986
                                                    Page 8

suggested  that  a  calculator be used, if available, to  reduce  the
chances of calculation error.

    Section 3.15.7 (Maintenance) provides the tester with a guide for
a maintenance program.  This program  is  not  required,  but  should
reduce equipment malfunctions.   Activity  matrices (Tables 5.1, 6.1,
and  7.1)  summarize all postsampling, calculation,  and  maintenance
activities.

5.  Auditing Procedure
    Section  3.15.8  (Auditing Procedures) provides a description  of
necessary activities for conducting performance  and  system  audits.
When  Method  7D  is  used  to  demonstrate compliance  with  an  EPA
pollutant emission standard, a performance audit must be conducted on
the  analytical  phase of the method.  The data processing procedures
and  a  checklist  for a systems audit  are  also  included  in  this
section.  Table 8.1 is an activity matrix for conducting the audits.

    Section   3.15.9   (Recommended   Standards    for   Establishing
Traceability) provides the primary standard  to  which  the  analysis
data should be traceable.  The primary  standard  is  sodium  nitrate
(NaNO3).

6.  References
    Section  3.15.10  contains  the promulgated  Method  7D;  Section
3.15.11  contains  the  references cited  throughout  the  text;  and
Section 3.15.12 contains copies of data  forms recommended for Method
7D.

-------
                                                Section No.  3.15
                                                Date July 1,  1986
                                                Page 9
                     PRETEST SAMPLING CHECKS
                     (Method 7D,  Figure 2.5)
Date 	  Calibrated by 	

Meter box number 	


Dry Gas Meter*

Pretest calibration factor (Y) = 	 (within 2% of
  average factor for each calibration run).


Rotameter

Pretest calibration factor (Y ) or setting = 	
  (between 400 and 500 cc/minl.
Dry Gas Meter Thermometer

Was a pretest temperature correction made? 	yes 	no

If yes, temperature correction 	 (within 3°C (5.4°F) of
  reference values for calibration and within 6 C (10.8°F) of
  reference values for calibration check).


Barometer

Was the pretest field barometer reading correct? 	yes 	no
 (within 2.5 mm (0.1 in.) Hg of mercury-in-glass barometer).
*Most significant items/parameters to be checked.

-------
                                                Section No. 3.15
                                                Date July 1, 1986
                                                Page 10
                       PRETEST PREPARATIONS
                     (Method 7D, Figure 3.1)
Apparatus check
Probe
Type liner
Glass
Stainless
steel
Other
Heated properly*
Leak checked
Filter
Glass wool
Other

Glassware
Restricted
orifice
impinger
Size
Type

Meter System
Leak- free pumps*
Rate meter*
Dry gas meter*
C02 Measurement
Orsat
Fyrite 	
Reagents
Water
Alkaline per-
manganate*
Silica gel
Other
Barometer
Drying tube
Acceptable
Yes







No







Quantity
required







Ready
Yes







No







Loaded
and packed
Yes







No







*Most significant items/parameters to be checked.

-------
                                                Section No.  3.15
                                                Date July 1,  1986
                                                Page 11
                         ON-SITE MEASUREMENTS
                       (Method 7D,  Figure 4.4)
Sampling

Impinger contents properly selected ,  measured,  and placed in
  impingers?*


Impinger Contents/Parameters*

1st: 200 ml of KMn04/NaOH 	

2nd: 200 ml of KMn04/NaOH 	

3rd: 200 ml of KMn04/NaOH 	
Drying tube: 6-16 mesh indicating type silica gel

Probe heat at proper level?* 	

Crushed ice around impingers? 	
Pretest leak check at 250 mm (10 in.) Hg?

Leakage rate? 	
Check of rotameter setting?
Probe placed at proper sampling point?
Flow rate constant at approximately 450 cc/min?*

C02 concentration measured?* 	
Posttest leak check at 250 mm (10 in.) Hg?*

Leakage rate?* 	
Sample Recovery

Contents of impingers placed in polyethylene bottles?

Fluid level marked?*
Sample containers sealed and identified?*
*Most significant items/parameters to be checked.

-------
                                                Section No.  3.15
                                                Date July 1,  1986
                                                Page 12
                       POSTTEST OPERATIONS
                     (Method 7D, Figure 5.4)
Reagents
Potassium  nitrate dried at 105 to 110 C for a minimum of 2 hours
before use? 	
Stock  standard  solution  (potassium nitrate) less than 2 months
old? 	

Sample Preparation
Has liquid level noticeably changed?* 	
  Original volume 	  Corrected volume

Analysis
Standard calibration curve prepared?* 	
Reagent blanks made from absorbing solution? 	
Same injection volume for both standards and samples? 	
Duplicate sample values agree within 5 percent of their mean?
Audit sample analytical results within 10 percent of true value?

All analytical data recorded on checklist and laboratory form?
* Most significant items/parameters to be checked.

-------
                                                   Section No. 3.15.1
                                                   Date July 1, 1986
                                                   Page 1
1.0 PROCUREMENT OF APPARATUS AND SUPPLIES
    The  procurement  of appropriate apparatus and  supplies  enables
quality results  to  be  obtained  from  Method  7D.   This   section
provides the user with information which complements the two sections
of Method  7D,  entitled "Apparatus" and "Reagents."  The information
is offered in the form of guidance and includes the following:

    o  Procedures for use in checking whether apparatus conforms with
       the requirements of the Method  7D  and corrective actions for
       when it does not  (Table  1.1  at  the  end  of  this  section
       summarizes  these procedures  and  also  contains  recommended
       corrective actions).

    o  Background  information  which can explain why specific appar-
       atus and reagents are required, and therefore, what limits may
       exist for alternatives or deviations.

    o  Practical information pertinent  to  the  selection and use of
       apparatus and reagents.

    o  Safety considerations.

    Persons responsible for the initial procurement  of apparatus and
supplies may find a procurement log helpful  in ensuring that all the
necessary  items are acquired and  are  in  good  working  order.   A
procurement log can  be  used  to record the descriptive title of the
equipment,  the  quantity, an identification number (if appropriate),
and the results  of acceptance checks.  An example procurement log is
shown  by  Figure  1.1,  a blank copy of this form  is  contained  in
Section 3.15.12 for the  Handbook  user.   Calibration  data obtained
during acceptance checks also should be recorded in a calibration log
book; see Section 2.0.

1.1  Sampling Apparatus

    Figure 1.2 shows the sampling train for  Method 7D.  It should be
noted that this sampling  train  is  very  similar  to  that used for
Method  6.   Several  of  the components and their use are identical,
including:

    o  Needle Valve
    o  Drying Tube
    o  Vacuum Pump
    o  Parts of the Metering System

This subsection  addresses  the specifications needed for procurement
purposes for all components  of  the  sampling  train  and associated
apparatus.

1.1.1 Sampling  Probe  - Method 7D specifies that sampling probes are
to be constructed of borosilicate glass.  The method also states that

-------
Item description
Qty.
Purchase

 order

 number
Vendor
                                     Date
Ord.
Rec.
Cost
Disposition
                                                                    Comments
      7Z4/3/
                         I/4/&
                    J/ZI/B+
                        •^J '
/i
                                Z//0/B4
                 Figure 1.1.   Example of a procurement log.
                                                                                      •d O W
                                                                                      0> 0) d>
                                                                                      (Q rt O
                                                                                      0) 0> rt
                                                                                          00
                                                                                        vO (-•
                                                                                        oo cn

-------
                                                    Section No. 3.15.1
                                                    Date  July 1, 1986
                                                    Page  3
PROBE END PACKED.
                                  RESTRICED ORIFICE IMPINGERS
                                                           SILICA GEL
                                                           DRYING TUBE
                Figure 1.2.    Method  7D.sampling  train.

-------
                                                   Section No. 3.15.1
                                                   Date July 1, 1986
                                                   Page 4

probes  made  of  either  stainless  steel or Teflon™ are acceptable.
Quartz probes (for example Vycor™)  may  be  used  for  sampling when
effluent temperatures exceed 480 C (900°F).

    The  function  of  the probe is rather  simple:  to  transport  a
representative effluent sample, cleaned of particulate matter, to the
impinger train.   To perform this function, the probe should:

   (a)   hold  a  filter  to  remove  particulate  matter,  including
         sulfuric acid mist;

   (b)   be constructed of a material that is unreactive toward NO ;
                                                                  Jv
   (c)   be free from leaks;

   (d)   be sufficiently long to enable samples  to  be acquired from
         the specified points(s) within the stack or duct;

   (e)   have  provisions  for  being  heated  in  order  to  prevent
         condensation of water vapor in the effluent sample; and

   (f)   be designed to connect to the inlet of the impinger train.

The three materials identified above  are unreactive toward NO .  The
appropriate  length  for  the  probe is determined primarily  Sy  its
intended  application  which will depend upon regulatory requirements
and the dimensions of the stack or duct where the measurements are to
be made.

   Sampling  probes  are  generally provided with electrical  heating
systems consisting  of  a  nichrome  wire which is wrapped around the
probe.   The  probe  and  heating system are, for protection,  placed
within a tightly  fitting tube, referred to as a sheath.  The heating
system should be capable of preventing condensation of water vapor in
the effluent sample stream  during  sampling.   Condensation  is  not
desired,   because  water  absorbs  N02  and  lowers   NO    results.
Additionally, if a stainless  steel  probe is used, condensation will
promote corrosion which shortens  probe  lifetime  and makes cleaning
difficult.

   It  is  recommended  that  probes  be  performance  checked before
initial use in the field to  ensure  that  condensation  can  be pre-
vented.  The  probe  should  first  be visually checked for cracks or
breaks  and  then  checked  for  leaks  according  to  the  procedure
described in Section 3.15.3 of this Handbook.  Then the probe heating
system should be checked as follows:

  1. Connect the probe (without filter) to the inlet of the pump.

  2.  Electrically  connect  and turn on the probe heater for 2 or  3
minutes.  If functioning properly, it will become warm to the touch.

-------
                                                   Section No. 3.15.1
                                                   Date July 1, 1986
                                                   Page 5

  3o  Start the pump, and adjust the  needle valve until a flow  rate
of between 400 and 500 cc/min is achieved.

  4=  Check the  probe.   It  should  remain  warm to the touch.  The
heater must be capable of maintaining the exit  air  temperature at a
minimum of 100°C (212 F) under these conditions.  If  it  cannot, the
probe should  be  rejected.   Any probe not satisfying the acceptance
check should be repaired, if possible, or returned to the supplier.

  The connection between  the  probe's  outlet and the impinger train
may  be  a  simple  fitting  or an additional length of tubing.   Any
connection should be leak-free.   In  addition, the connection should
be constructed of borosilicate glass,  stainless steel, or Teflon? and
therefore,  like  the  probe be unreactive toward  NO .   Lastly,  if
tubing is used,  provisions  should  exist to prevent condensation of
water vapor upstream  of  the  impinger  train  during  sampling.   A
heating system for the connection  is  not  required  if  the probe's
heating system can supply enough heat to the effluent sample.

1.1.2  Restricted Orifice Impingers - The sampling train requires the
use  of  three  restricted orifice  impingers  connected  in  series.
Figure  1.3  shows  one  of  these impingers, which are  commercially
available.

                                                              13
  Impinger design is  important  to  obtain  quality  results.    The
restricted  orifice  impinger is specifically designed to promote the
complete  collection  of  NO,  which  is relatively unreactive.   Two
design features are important; (a) the length of  the  liquid column,
and  (b) the size of the  impinger's stem tip.  The impingers used for
Method 7D are narrower than  Greenburg-Smith  impingers  in  order to
provide a greater  depth of absorbing reagent and, hence, to increase
the  reaction  time  of  the sample gas  in  the  absorbing  reagent.
Because of the narrow opening of the stem tip, the effluent sample is
introduced into the absorbing  reagent  as  smaller bubbles.  Smaller
bubbles  promote the reaction of NO because of their greater surface-
to-volume ratio and thus, greater exposure to the absorbing reagent.

  Impingers  with  stem  tips restricted to less than 1.5 mm internal
diameter  are  easily  plugged  by  reaction products.   The  problem
typically  affects  only  the  first impinger of the  sampling   train
because:  (a) most of the NO  and C02, and (b) all the sulfur dioxide,
if   present, are reacted there.  If plugging occurs, the problem  may
be minimized by making  the  length  of the capillary tubing shorter;
plugging also can be minimized  by keeping stem tips clean.  Reaction
products in the stem  tips  can  be  removed by immersion in either 3
percent by  volume hydrogen peroxide solution  [3% (v/v) H909  (aq)] or
3M hydrochloric acid solution  [HC1  (aq)].  CAUTION:  Chlorine  (Cl..,)
gas  is  evolved  during the use  of  the  HC1;  therefore,  cleaning
operations should be conducted in a fume hood.  The H.-02 solution  is
identical to the absorbing solution used for Method 6.

-------
         45/50
DIMENSIONS: mm
         35

     CAPILLARY
     TUBING:
     1.5 I.D.
                                            Section No. 3.15.1
                                            Date  July 1, 1986
                                            Page  6
                              28/12
      Figure 1.3.  Restricted orifice impinger.

-------
                                                   Section No. 3.15.1
                                                   Date July 1, 1986
                                                   Page 7

    It is recommended that each impinger upon receipt be checked vis-
ually for damage,  such  as  breaks  or cracks, and for manufacturing
flaws, such as poorly shaped connections.

    Other nonspecified collection absorbers  and  sampling flow rates
may  be  used,  subject  to  the  approval to the Administrator,  but
collection efficiency  must  be  shown to be at least 99% for each of
three test  runs  and must be documented in the emission test report.
For efficiency testing,  an extra absorber must be added and analyzed
separately and must not contain more than 1% of the total NO .
                                                            a

1.1.3  Vacuum Pump - The vacuum pump should be capable of maintaining
a flow rate of approximately 400 to 500 cc/min for pump inlet vacuums
up to 250 mm (10 in.) Hg with the pump outlet near standard pressure,
[i.e., 760 mm (29.92  in.)  Hg].   The  pump  must  be leak free when
running  and  pulling a vacuum (inlet plugged) of 250 mm (10 in.) Hg.
Two  types  of  vacuum pumps are commonly  used:  either  a  modified
sliding fiber vane pump or a diaphragm pump.  For safety reasons, the
pump should be equipped with a three-wire electrical cord.

    To check the pump for leaks, install  a  vacuum gauge in the pump
inlet line.  Plug the inlet line,  and  run the pump until the vacuum
gauge reads 250 mm (10 in.)  Hg  of  vacuum.   Clamp  the pump outlet
line, and turn off the  pump; the vacuum reading should remain stable
for 30 seconds.

1.1.4  Volume Meter  - The dry gas meter must be capable of measuring
total  volume  with  an  accuracy  to  within 2%, calibrated  at  the
selected flow rate (between 400 and 500 cc/min), and must be equipped
with a temperature gauge (dial thermometer, or eguivalent) capable of
measuring the gas temperature to within 3°C (5.4 F).

    A  new dry gas meter may be checked for damage  visually  and  by
performing  a  calibration  according  to Section 3.5.2 of this Hand-
book.  Any dry gas meter  that  is  damaged,  behaves erratically, or
does not give readings within 4-2% of the selected flow rate for  each
calibration  run is unsatisfactory.  Also  upon  receipt,  the  meter
should be calibrated  over  a varying flow range to see whether there
is any effect on the calibration.

    Dry gas meters that  are  equipped  with temperature compensation
must  be  calibrated  over the entire range of temperatures that  the
meter encounters under actual field conditions.  The calibration must
contain  at  least  one  data  point  at  each  10°F  interval.   All
temperatures that are to be used in the field  must  be  within 2% of
the calibrated value.

    The  wet  test  meter used to check the dry gas meter  should  be
calibrated  using  the primary displacement  technique  explained  in
Section 3.5.2 of this Handbook.   The  wet3 test  meter  must  have a
capacity of at least 0.0003 m /min (0.1 ft /min) with an accuracy  of
+2%; otherwise at the higher flow rates,  the water will not be level
and this possibly will result in an incorrect reading.

-------
                                                   Section No. 3.15.1
                                                   Date July 1, 1986
                                                   Page 8

1.1.5  Rotameter - A rotameter, or its equivalent, with a range of  0
to 1 L/min isused to monitor the sampling flow rate.  The rotameter
is checked against the calibrated  dry  gas meter with which it is to
be  used  or against a wet test meter.  It is recommended that it  be
within 5%  of  the  manufacturer's  calibration curve.  The rotameter
flow setting of about 450 cc/min should be determined.

    Changes in pressure, density, and  viscosity  of  the  sample gas
will affect the calibrated  sample rate.   However, since sampling is
performed  at  a constant rate, which need not be  isokinetic,  these
changes  do  not  affect  the  sample  volume measured by the dry gas
meter.

1.1.6   Needle Valve -  A  metering  valve  with  conveniently  sized
fittings is required in  the sampling train to adjust and control the
sample flow rate.  It is recommended that the needle valve  be placed
on the vacuum side of the pump.

1.1.7   Drying  Tube - The drying tube should be packed  with  6-  to
16-mesh  indicating-type silica gel, or equivalent, to dry the sample
gas and to protect the pump and the meter.  A drying tube can be made
by  filling a 10-mm polyethylene tube with  silica  gel  and  packing
glass  wool in each end to hold  the  silicia  gel  and  protect  the
sampling system.  Plastic tubing can  be  utilized in any connections
downstream  of the impinger  train  without  affecting  the  sampling
results.  The drying tube should  have a minimum capacity of 30 to 50
g of silica gel and should  be visually checked before use for proper
size and for damage.

    If  the silica gel has been used previously, it must be dried  at
175°C  (350 F) for 2 hours.  New silica gel may be used as  received.
Other types of desiccants  may  be  used  subject  to approval of the
Administrator.

1.1.8   Metering System - For ease of use, the metering system—which
contains  the  dry gas meter, thermometer(s),  vacuum  pump,   needle
valve, and rotameter--can be assembled into  one  unit  (meter  box).
After  a  meter  box  has  been either constructed or purchased, then
positive and negative pressure leak checks should be performed.

    The positive  pressure  leak  check,  similar  to  the  procedure
described in Method 5 (Section 3.4) of this Handbook, is performed as
follows:

    1.   Attach  rubber tubing and inclined manometer,  as  shown  in
Figure 2.1 of Section 3.4.2 of this Handbook.

    2.  Shut off the needle valve, and apply positive pressure to the
system by blowing into the rubber tubing until the inclined manometer
or magnehelic gauge reads from 12.5 to 17.5 cm (5 to   7 in.) H0.

-------
                                                   Section No. 3.15.1
                                                   Date July 1, 1986
                                                   Page 9

    3.  Pinch off the tube, and observe  the  manometer for 1 minute.
A  loss of pressure indicates a leak of the apparatus  in  the  meter
box.

    After the meter box apparatus has passed the positive leak check,
then the negative leak check should be performed as follows:

    1.  Attach the vacuum gauge at the inlet to  the drying tube, and
pull a 250 mm (10 in.) Hg vacuum.

    2.   Pinch  or clamp the outlet of the flow meter.  This  can  be
accomplished by closing the optional shutoff valve if employed.

    3.  Turn off the  pump.   Any  deflection  noted  in  the  vacuum
reading within 30 seconds indicates a leak.

    4.   Carefully release the vacuum gauge before releasing the flow
meter end.

    If either of these checks detects  a  leak  that  cannot  be cor-
rected,  the  meter  box  must  be rejected and/or  returned  to  the
manufacturer.

    The dry gas meter must be equipped with a temperature gauge  (dial
thermometer or equivalent).  It is recommended that upon receipt this
be checked visually for damage,  such  as  dents or a bent stem.  The
thermometer  should  read  within  3 C (5.4 F) of the true value when
checked   at   two   different   ambient   temperatures   against   a
mercury-in-glass thermometer  that  conforms  to  ASTM E-l No. 63C or
63F.  The two ambient temperatures  used to calibrate the thermometer
must  differ  by a minimum of 10 C (18 F).  Damaged thermometers that
cannot be calibrated are to be rejected.

    1.1.9  Barometer - A mercury, aneroid, or other barometer capable
of measuring atmospheric pressure to within  2.5  mm  (0.1 in.) Hg may
be  used.   However, in many cases, the barometric  pressure  can  be
obtained from  a  nearby  National  Weather Service Station, in which
case the  station  value   (which is the absolute barometric pressure)
should be requested.  The tester should be aware that the pressure is
normally corrected to sea level by the  weather  station;  the uncor-
rected readings should be obtained.  An adjustment for differences in
elevation of the weather station and the sampling location is applied
at  a  rate  of  -2.5  mm  Hg/30  m (-0.1 in. Hg/100  ft) of elevation
increase, or vice versa for elevation decrease.

    Accuracy can be ensured by checking the field barometer against a
mercury-in-glass barometer or its equivalent.  If the field barometer
cannot be  adjusted  to agree with the mercury-in-glass barometer, it
is not acceptable.

1.1.10   Vacuum  Gauge - At least one 760-mm (29.92-in.) Hg gauge  is
necessary  to  leak  check the sampling train.  An acceptable  vacuum

-------
                                                   Section No. 3.15.1
                                                   Date July 1, 1986
                                                   Page 10

gauge, when checked in a parallel leakless  system  with a mercury U-
tube manometer at 250-mm (10-in.)  Hg vacuum, will agree within 25 mm
(1.0 in.) Hg.

1.2  Sample Recovery Apparatus

1.2.1  Wash Bottles - Two 500-ml polyethylene  or  glass wash bottles
are needed for quantitative recovery of collected samples.

1.2.2  Storage  Bottles  - One 1-L polyethylene bottle is required to
store  each  collected sample.  An additional polyethylene bottle  is
necessary  to  retain  a  blank  for  each absorbing solution used in
testing.  Wash and storage bottles should  be  visually  checked  for
damage.  CAUTION:  Each storage bottle  seal  should be checked prior
to use to ensure that leakage will not occur.

1.2.3  Funnel and Stirring Rods - The analyst may find a glass funnel
and  glass  stirring  rods  are helpful in transferring the absorbing
reagent to and from the restricted  orifice  impingers.   The flow of
absorbing  reagent  can  be  controlled by pouring  along  the  glass
stirring rod.

1.3  Apparatus for Sample Preparation and Analysis

1.3.1  Magnetic Stirrer with Magnetic Stirring  Bars  -  The magnetic
stirrer  and  stirring  bars  are  used  for  the removal  of  excess
permanganate  ion.   The stirring bars should be Teflon™-coated owing
to the corrosiveness of the alkaline-permanganate solution.  Stirring
bars  having  dimensions  25  mm  by  10 mm are recommended.  Smaller
stirring bars can be expected  to  be  less  efficient because of the
resistance  offered  by  the  absorbing reagent, which is  relatively
viscous.

    Manual stirring is acceptable; however, because it is tedious and
laborious, this is not recommended.

1.3.2  Filtering Flask - One filtering flask having a 500-ml capacity
is  needed  to filter the liquid sample after the excess permanganate
ion has been removed.

1.3.3  Buchner Funnel - The Buchner funnel is used with the filtering
flask for the filtering operations.  A convenient  size funnel is one
with  a  75-mm  internal diameter.  The analyst may wish to attach  a
section of Teflon™tubing  to  the  funnel's spout in order to prevent
loss of sample via the side arm of the flask during filtration.   The
use of a trap  located  between  the  filtering  flask and the vacuum
source  is  also recommended in order to prevent  inadvertant  sample
losses.  Upstream tubing connections for the trap should be Teflon™.

1.3.4   Filter  - Whatman GF/C glass microfiber discs are used in the
Buchner  funnel.   For  the  funnel  size  recommended   above,   the \

-------
                                                   Section No.  3.15.1
                                                   Date July 1, 1986
                                                   Page 11

applicable disc diameter is 7.0 cm.  This filter is specified because
it  performs  well  on materials having a small particle  size.   The
material  filtered  from  the  sample is primarily manganese  dioxide
which exists in very small particles when suspended in water.

    In general, other types of filters are unsuitable owing either to
clogging  or to their  inability  to  filter  the  manganese  dioxide
particles  effectively.   The  analyst  should note that  particulate
matter  must  be  removed from the sample in order to avoid damage to
syringes and the ion chromatograph.

1.3.5  Vacuum  Source  -  A vacuum source is needed for the filtering
operations.  Either a water aspiration system or a vacuum pump can be
used.  If a  vacuum  pump  is  used, it should be protected by a trap
installed at an upstream position.

1^3.6  Funnel and Stirring  Rods  - The analyst may find a funnel and
glass stirring rods are helpful in transfering  the sample aliquot to
the Erlenmeyer flask prior to removal of the excess permanganate ion.

1.3.7  Volumetric Flask - One  volumetric  flask  having  the Class-A
designation and a 250-ml capacity  is  needed  for  each  sample  and
blank.  As a practical matter, samples should be stored in the flasks
for_ a  minimum  period of time owing to the fact that hydroxide ions
(OH~) will  attack  the  glass and can also cause frozen ground-glass
fittings.

1.3.8   Pipettes  - A 50-ml Class-A pipette is needed  for  taking  a
sample  aliquot.   A  5-ml  pipet  is  usually  used for adding  (not
quantitatively) hydrogen peroxide  to  the sample aliquot in order to
remove excess permanganate ion.  Because hydroxide  ion  (OH~), which
is present in the sample,  can  attack  glass, it is recommended that
analysts rinse pipettes with water immediately after  use on samples.
If  Quality  Assurance  Audit  Samples are to be analyzed, additional
pipettes (Class-A) may be needed.

1.3.9  Erlenmeyer Flasks - Erlenmeyer flasks having a 250-ml capacity
are used for operations involving  the removal of excess permanganate
ion in the samples.

1.3.10  Ion Chromatograph - An  ion  chromatograph  (1C)  is used for
analyzing the samples.  The instrument should, at a minimum, have the
components described below.

    Columns  - The 1C should be equipped with an ion separator column
capable  of  resolving  nitrate  ion  (N03~) from sulfate ion (S04=),
which may be  found  in  samples  acquired at fossil-fuel-fired steam
generators.  In addition, it  should  be  capable  of  detecting  and
resolving nitrite ion (NO-')-  Either  suppressed   or  nonsuppressed
IC's may be used provided  that performance meets the above criteria.
Suppressed  IC's  should  be  equipped  with  an acid (H ) suppressor
column in addition to the anion separator column.  Suppressor columns

-------
                                                   Section No. 3.15.1
                                                   Date July 1, 1986
                                                   Page 12

(fiber  preferred  over  packed  bed)  are  generally   produced   as
proprietory items; however, an acceptable  column  can  be made using
the resin available from BioRad Company,  32nd  and  Griffin Streets,
Richmond, California.

    Pump  -  The pump must be capable of maintaining a steady  eluent
flow as required by the system.

    Flow Gauges - These must  be  capable  of measuring the specified
eluent flow rate.  It is recommended  that  the  gauge  be calibrated
upon receipt.

    Conductivity Detector  -  It  is recommended that the detector be
calibrated  according  to  manufacturer's procedures prior to initial
use.

    Recorder - It should be compatible with the output voltage of the
detector.

1.3.11  Analytical Balance - One analytical  balance  that  weighs to
0.1 mg and a set of Class-S calibration weights to check the accuracy
of the balance (+0.3 mg) upon receipt are needed.  The balance should
be serviced or returned to the manufacturer  if  agreement  cannot be
met.

1.4  Reagents - Unless otherwise indicated, it is intended  that  all
reagents  conform  to the specifications established by the Committee
on Analytical  Reagents of the American Chemical Society (ACS), where
such  specifications  are  available;  otherwise,  use the best grade
available.

1.4.1  Sampling - For sampling, the following are needed.

    Absorbing solution - The absorbing  solution  is prepared by dis-
solving 40.0 g potassium  permanganate  (KMnO.)  and  20.0  g  sodium
hydroxide (NaOH) in 940 ml of water.  The solution's concentration is
4.0  percent  (w/w)  KMnO., 2.0 percent (w/w) NaOH.  CAUTION: Extreme
care should be taken in  Handling the KMnO, reagent and the absorbing
solution.   KMnO.  is  a  strong  oxidant  and  is  incompatible with
substances  containing  carbon  such  as  paper,  fabric,  and  human
tissue.  It is recommended that  eye protection be worn when handling
the  absorbing  solution.  Skin exposed  to  the  absorbing  solution
should be washed with plenty of water and  until  the exposed area no
longer exhibits a soapy feeling.

    Water - Water should  be used which conforms with ASTM specifica-
tion D1193-82, Type III.  Type III water is prepared by distillation,
ion exchange, reverse osmosis, or a combination thereof, followed  by
polishing  with  a  0.45 ym membrane filter.  The specifications  for
Type III water are shown below.

-------
                                                   Section No.  3.15.1
                                                   Date July 1, 1986
                                                   Page 13

          Specifications for ASTM D1193 - 82,  Type III Water

          Total matter, max., (mg/L)            1.0

          Electrical conductivity, max.,        1.0
           (umho/cm) at 25 C

          Electrical resistivity, min.,         1.0
           (umho/cm) at 25 C

          pH at 25°C                        6.2 to 7.5

          Minimum color retention time          10
           of KMn04, (min)

          Maximum soluble silica, (yg/L)        10

Note: Mention of "water" anywhere  in  this  Section (3.15) refers to
ASTM  D1193-82,  Type  III Water as described above.  By using  water
from the same source for making reagents, calibration standards,  and
eluents for the ion chromatograph, the presence  of  trace quantities
of nitrate in the water will be  negated.   Therefore,  a water blank
correction is not necessary  in  the  development  of the calibration
curve.

    Stopcock  Grease - An acetone insoluble,  heat  stable,  silicone
grease must  be  used  when  sealing  of  ground-glass connections is
required.

1.4.2  Analysis - For analysis, the following reagents are required.

  Water - See Subsection 1.4.1 above.

  Hydrogen Peroxide - Five (5) percent  (v/v) hydrogen peroxide  (H_O2)
is used  which  is  prepared  by  mixing 1 part 30% (v/v) H202 with 5
parts water.

  Reagent Blank - The reagent blank  may  be  prepared  by dissolving
2.4 g KMnO. and 1.2 g NaOH in 96 ml water.  Alternatively,  the blank
may be prepared  by diluting 60 ml of the absorbing reagent to  100 ml
using water.

  Potassium  Nitrate   (KNO^)  Standard  Solution   -   The  following
procedure is observed to prepare the KNO~ standard solution.

   1. Dry an adequate amount of KN03 at 110°C for about 2 hours; then
transfer   to   a   desiccator,  ana  allow  to  cool  to  laboratory
temperature.

   2. Using  an analytical balance, accurately weigh 9 to 10 g of the
dried KN03 to the nearest 0.1 mg.

-------
                                                   Section No. 3.15.1
                                                   Date July 1, 1986
                                                   Page 14

   3. Transfer  the  KN03  to a suitable container, such as a beaker,
dissolve the KN03 in water, and transfer all of the  KN03 solution to
a 1-L volumetric flask.

   4. Dilute the KN03 solution to the 1-L mark with water.

The  N03~  concentration  of the standard solution is calculated from
the mass of KN03 using the following relationship:

    NO ~         /Mass of KNOJV/106 vg\/   L  W 62.01 g/mol NO " \
Concentration   =\   (g)      )\   g  /\1Q3 mlMl01.10 g/mol KNO3 f
  (yg/ml)

Method  7D  states  that  the  KN03 standard solution is stable for 2
months   without  preservative  at  laboratory  conditions.    Novice
analysts  should  note  that certain microbes feed on N03   solutions
with the consequence for Method  7D  being  that  NO  results will be
biased high.  For this reason,  standard solutions snould be disposed
of after 2 months.

    Eluent Solution - For IC's involving the suppressedgtechnique, an
eluent solution being 3 x 10   M NaHC03  and  2.4 x 10   M Na2C03 has
proved adequate for  Method 7D applications.  This eluent is prepared
by  taking  1.008  g NaHC03 and 1.018 g Na2C03 and dissolving them in
4 L water.

    Other eluents may  be  used  provided  that  they  are capable of
resolving N0~~ from SO." and other  ions  which  may  be  present  in
samples.

    Quality Assurance Audit Samples - Quality Assurance Audit Samples
are required to be analyzed in conjunction  with  field  samples when
Method  7D  is  used  to demonstrate compliance with EPA's New Source
Performance  Standards  in  40  CFR  Part  60.  The audit samples for
Method 7D are essentially the same as those  described  in  Method 7,
Section 3.3.9.  Because the analytical range  for  Method  7D differs
from  that  for  Method 7, analysts requesting audit  samples  should
specify that samples be appropriate for Method 7D.

-------
                                                                   Section No.  3-15-1
                                                                   Date July 1, 1986
                                                                   Page 15
        Table 1.1.  ACTIVITY MATRIX FOR PROCUREMENT OF APPARATUS AND SUPPLIES
Apparatus and
  supplies
  Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Sampling probe
 with heating
 system
Capable of maintaining
100°C (212°F) exit
air at flow rate of
500 cc/min
Visually check and
run heating system
checkout
Repair, or
return to
supplier
Restricted
 orifice
 impingers
Standard stock glass;
ensure that dimensions
conform with specifi-
cations
Visually check upon
receipt for breaks
or leaks
Return to manu-
facturer
Vacuum pump
Capable of maintaining
flow rate of 400 to
500 cc/min; leak free
at 250 mm (10 in.) Hg
Check upon receipt
for leaks and capacity
As above
Dry gas meter
Capable of measuring
total volume within
2% at a flow rate of
500 cc/min
Check for damage upon
receipt, and calibrate
(Sec. 3.15.2) against
wet test meter
Reject if dam-
aged , behaves
erratically,
or cannot be
properly
adjusted
Wet test meter
Capable of measuring
total volume within
2% at a flow rate of
500 cc/min
Upon assembly, leak
check all connections,
and check calibration
by liquid displacement
As above
Rotameter
Within 5% of manufac-
turer's calibration
curve (recommended)
Check upon receipt for
damage, and calibrate
(Sec. 3.15.2) against
wet test meter
Recalibrate,
and construct
a new calibra-
tion curve
Drying tube
Minimum capacity of
30 to 50 g of silica
gel
Visually check upon
receipt for damage and
proper size
Return to
supplier
(continued)

-------
Table 1.1 (continued)
                                                                   Section No. 3.15.1
                                                                   Date July 1, 1986
                                                                   Page 16
Apparatus and
  supplies
   Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Thermometers
Within 1°C (2°F)
true value in the range
of 0°C to 25°C
(32° to 77°F)
for impinger and within
3°C (5^°F) for
dry gas meter thermom-
eter
Check upon receipt for
damage (i.e., dents and
bent stem),  and
calibrate (Sec. 3.15.2)
against mercury-in-
glass thermometer
Return to
supplier if
unable to
calibrate
Barometer
Capable of measuring
atmospheric pressure
to within 2.5 mm
(0.1 in.) Hg calibrate
Check against mercury-
in-glass barometer or
equivalent (Sec. 3.5-2)
Determine cor-
rection factor,
or reject if
difference is
more than
2.5 mm
Vacuum gauge
Wash bottles
Storage
 bottles
Pipettes and
 volumetric
 flasks
0 to 760 mm (0 to
29.92 in.) Hg range,
^25 mm (1.0 in.) Hg
accuracy at 250 mm
(10 in.) Hg
Check against U-tube
mercury manometer
upon receipt
Adjust, or re-
turn to
supplier
Polyethylene or glass,
500-ml
Visually check for
damage upon receipt
Replace, or re-
turn to
supplier
Polyethylene, 1-L
Visually check for dam-
age upon receipt, and
be sure that caps seal
properly
As above
Glass, Class-A
Upon receipt, check for
stock number, cracks,
breaks, and manufac-
turer flaws
As above
Water
Must conform to ASTM-
D1193-82, Type III
Check each lot or spec-
ify type when ordering
As above
 (continued)

-------
Table 1.1 (continued)
                                                                   Section No.  3.15.1
                                                                   Date July 1, 1986
                                                                   Page 17
Apparatus and
  supplies
  Acceptance limits
Frequency and method
    of measurement
  Action if
  requirements
  are not met
Stopcock
 grease
High vacuum, high temp-
erature chlorofluoro-
carbon grease
Visually check upon
receipt
Return to
supplier, and
note in procure-
ment log
Analytical
 balance
Capable of measuring
to +0.1 mg
Check with standard
weights upon receipt
and before each use
Replace, or
return to manu-
facturer
Ion Chroma-
 tograph
  1.  Columns
  2.  Pump
  3.  Flow
      gauges
1.  Capable of giving
nitrate ion peaks with
baseline separation
1.  Check during
analyses
2.  Capable of deliv-
ering eluent at con-
stant and repeatable
flow rate
2.  Check during
analyses by monitor-
ing flow rate
3.  Capable of giving
repeatable indications
of eluent flow rate
3.  Check calibration
and repeatability
upon receipt
  1.  Consult op-
erators '  manual;
regenerate sup-
pressor column;
clean separator
column; check
performance of
components below;
replace column(s)
if above actions
are unsuccessful

  2.  Consult op-
erator's manual;
oil, clean, re-
repair, replace,
or return to man-
ufacturer; check
tubing of ion
chromatograph for
leaks or ob-
structions;
check flow meter
performance

3.  Consult oper-
ator 's manual;
adjust, repair,
replace,  or
return to
manufacturer
 (continued)

-------
Table 1.1 (continued)
                                                                   Section No. 3.15.1
                                                                   Date July 1, 1986
                                                                   Page 18
Apparatus and
supplies
4 . Conduc-
tivity
detector
5 . Recorder
Hydrogen per-
oxide
Potassium
nitrate
Sodium carbonate
Sodium bicarbon-
ate
Sodium hydroxide
Potassium
permanganate
Acceptance limits
4. Capable of giving
responses which can be
manually or electron-
ically integrated
within a precision of
5 percent
5. As above, if used
record responses for
manual integration
30/li aqueous solution,
ACS reagent grade
ACS reagent grade
ACS reagent grade
ACS reagent grade
ACS reagent grade
ACS reagent grade
Frequency and method
of measurement
4. Calibrate accord-
ing to manufacturer's
instructions prior to
use
5. Check during
analyses
Check each lot, or
specify type when
ordering
As above
As above
As above
As above
As above
Action if
requirements
are not met
4. Consult
operator ' s
manual ; repair
replace, or
return to
manufacturer
5. Consult
operator's
manual; adjust
speed
Replace or
return to
manufacturer
As above
As above
As above
As above
As above

-------
                                                   Section No.  3.15.2
                                                   Date July 1,  1986
                                                   Page 1
2.0  CALIBRATION OF APPARATUS
    Calibration of  the  apparatus  is  one  of  the  most  important
functions  in  maintaining  data quality.  The  detailed  calibration
procedures included in this section were  designed  for the equipment
specified  in Method 7D and described in the previous section.  Table
2.1  at  the  end of this section summarizes  the  quality  assurance
functions  for  calibration.   All calibrations should be recorded on
standardized forms and retained in a calibration log book.

    The calibration procedures and considerations  addressed  in this
section are those which are unique to Method 7D.  The sampling  phase
of  Method  7D  involves the use of equipment that is essentially the
same as that used for Method  6.   The  analysis  phase  of Method 7D
entails the use  of  an ion chromatograph,  an instrument that also is
used  for  Method  7A.  The Handbook user should note that:  (a)  the
standard  used  for Method 7A is sodium  nitrate  (NaN03,  while  for
Method  7D  the standard used is potassium nitrate  (KNo3);  and  (b)
sulfate ion (SO,") peaks in ion chromatograms for Method 7D will have
a lesser tendancy to overlap and therefore  to interfere with nitrate
(N03~) peaks because SO." will exist at a lower concentration because
it originates only from sulfur oxides in the effluent.

2.1  Metering System

2.1.1.  Wet Test Meter - The wet  test  meter  must be calibrated and
have  the proper capacity.  For Method 7D,  the wet test meter  should
have a capacity of at least 1 L/min.  No upper limit is placed on the
capacity; however,  the  wet test meter dial should make at least one
complete  revolution at the specified flow rate for each of the three
independent calibrations.

    Wet test meters are calibrated by the manufacturer to an accuracy
of +2%.  Calibration  of the wet test meter must be checked initially
upon receipt and yearly thereafter.

    The following liquid  positive displacement technique can be used
to  verify  and  adjust, if necessary, the accuracy of the  wet  test
meter to +2%:

    1.  Level the wet test  meter  by  adjusting  the  legs until the
bubble on the level located on the top of the meter is centered.

    2.  Adjust the water volume in the  meter  so that the pointer in
the water level gauge just touches the meniscus.

    3.  Adjust the water manometer to  zero by moving the scale or by
adding water to the manometer.

    4.  Set  up  the apparatus and calibration  system  as  shown  in
Figure 2.1.

-------
                                               Section No. 3.15.2
                                               Date  July 1, 1986
                                               Page  2
                MANOMETER
      THERMOMETER
AIR INLET
                                   WATER
                                   LEVEL
                                   GAUGE
                                                      VALVE
                                                      2000-ml LINE
                                                         TYPE-A
                                                         VOLUMETRIC
                                                         FLASK
Figure  2.1.   Calibration check  apparatus for  wet test meter.

-------
                                                   Section No.  3.15.2
                                                   Date July 1,  1986
                                                   Page 3

        a.    Fill the rigid-wall 5-gal jug with water  to  below  the
             air inlet tube.   Put water in the impinger or saturator,
             and allow both to equilibrate to room temperature (about
             24 hours) before use.

        b.    Start  water  siphoning  through the system, and collect
             the water  in a 1-gal container, located in place of the
             volumetric flask.

    5.   Check operation of the meter as follows:

        a.    If the manometer is reading <10  mm  (0.4  in.) H-O, the
             meter is in proper  working condition.  Continue to step
             6.

        b.    If  the manometer reading is >10 mm (0.4 in.)  H.-0,  the
             wet test meter is defective.  If  the  wet test meter is
             defective, and if the defects(s) (e.g., bad  connections
             or joints) cannot be found and corrected,  return  it to
             the manufacturer for repair.

    6.   Continue the operation until the 1-gal  container  is  almost
full.  Plug the inlet  to the saturator.  If no leak exists, the flow
of  liquid  to  the  gallon  container  should  stop.   If  the  flow
continues,  correct for  leaks.  Turn the siphon system off by closing
the valve,  and unplug the inlet to the wet test meter.

    7.   Read  the  initial volume (V.) from the wet test meter  dial,
and record on the wet test meter calibration log, Figure 2.2.

    8.   Place  a  clean, dry volumetric  flask  (Class-A)  under  the
siphon tube, open the pinch clamp, and  fill  the volumetric flask to
the mark.  The volumetric  flask  must  be  large  enough to allow at
least  one  complete  revolution  of the wet test meter with not more
than two fillings of the volumetric flask.

    9.   Start  the  flow  of  water,  and record the maximum wet test
meter manometer reading  during  the  test  after  a constant flow of
liquid is obtained.

   10.   Carefully fill _the volumetric flask,  and shut off the liquid
flow at the 2-L mark.  Record  the final volume shown on the wet test
meter.

   11.   Steps 7 through. 10 must be performed three times.

    Since the water temperature in the  wet  test meter and reservoir
has been equilibrated to the ambient  temperature and the pressure in
the  wet  test  meter will equilibrate with the water reservoir after
the  water  flow is shut off, the air volume can be compared directly
with  the liquid displacement volume.  Any  temperature  or  pressure

-------
    test meter serial number
                                4 5 ~~
Date  2-/2.1/&4-
  Satisfactory  leak check? 	

                               .                                                  ~7sf ^ r~
  Ambient  temperature of equilibrate liquid in wet test meter and  reservoir    /y- /""
Range of wet  test  meter flow rate 0~~12-0 (—/^ir\

Volume of  test  flask Vg =  -2- > 00 L-	
Test
number
1
2
3
Manometer
reading, a
mm H2O
r
5~
       ^~m    s   s
                                               (+1%).
                                 Signature of calibration person
                                                                                                        V O OT
                                                                                                        o>a>a>
                                                                                                        (Q ft O
                                                                                                        0) 0> rt
                                                                                                            H-
                                                                                                        t>-<-iO
                                                                                                          C 3
                         Figure 2.2.  Wet test meter calibration log.
                                                                                                            w
                                                                                                          M •
                                                                                                          VD (-•
                                                                                                          oo cn
                                                                                                          CT» •
                                                                                                            to

-------
                                                   Section No. 3.15.2
                                                   Date July 1, 1986
                                                   Page 5

difference would be less than measurement  error and would not affect
the final calculations.
    The  error  should  not exceed +^1%; if this  error  magnitude  is
exceeded,  check all connections within the test apparatus for leaks,
and gravimetrically check the volume  of  the standard flask.  Repeat
the calibration procedure, and if the tolerance  level  is  not  met,
adjust  the liquid level within the  meter  (see  the  manufacturer's
manual) until the specifications are met.

2.1.2   Sample  Metering  System - The sample metering  system,  con-
sisting of the drying tube, needle valve,  pump,  rotameter,  and dry
gas meter, is initially  calibrated  by  stringent laboratory methods
before  it  is  used in the field.  The calibration is then rechecked
after each field test series.  This recheck requires less effort than
the initial calibration.  When a recheck indicates that the  calibra-
tion factor  has  changed, the tester must again perform the complete
laboratory procedure to obtain the new calibration factor.  After the
meter is recalibrated, the metered sample volume is multiplied by the
calibration factor (initial or recalibrated)  that  yields  the lower
gas volume for each test run.

    Initial Calibration - The metering  system  should  be calibrated
when first  purchased  and  at  any  time the posttest check yields a
calibration factor that does not  agree  within  5%  of  the  pretest
calibration factor.   A  calibrated  wet  test meter (properly sized,
with +1% accuracy) should be used to calibrate the metering system.

    The metering system should be calibrated in the following  manner
before its initial use in the field.

    1.  Leak  check  the  metering system (drying tube, needle valve,
pump, rotameter, and dry gas meter) as follows:

        a.   Temporarily attach a suitable rotameter (e.g., 0-40 cm /
             min)  to  the  outlet of the dry gas meter, and place  a
             vacuum gauge at the inlet to the drying tube.

        b.   Plug the drying tube  inlet.   Pull a vacuum of at least
             250 mm (10 in. ) Hg.

        c.   Note the flow rate as indicated by the rotameter.

        d.   A  leak of <0.02 L/min must be recorded or leaks must be
             eliminated.

        e.   Carefully  release  the vacuum gauge before turning  off
             pump.

    2.  Assemble  the apparatus, as shown in Figure 2.3, with the wet
test  meter replacing the drying tube and impingers; that is, connect
the outlet of  the  wet  test  meter  to the inlet side of the needle

-------
THERMOMETER
                                                                  MANOMETER
                                                                                AIR INLET
Figure 2.3.   Sample metering system calibration  setup.
                                                                                              •a o OT
                                                                                              0) 0) (D
                                                                                              (Q rt o
                                                                                              a> a> rt
                                                                                                  H-
                                                                                              a\ Q o
                                                                                                c 3
                                                                                               VD H"
                                                                                               00 CJ1
                                                                                               CT» •
                                                                                                 to

-------
                                                   Section No. 3.15.2
                                                   Date July 1, 1986
                                                   Page 7

valve and the inlet side of the wet test meter  to  a saturator which
is open to the atmosphere.  Note: Do not use a drying tube.

    3.  Run  the  pump  for  15  minutes  with  the  flow rate set at
450 cc/min to allow  the  pump  to warm up and to permit the interior
surface of the wet test meter to become wet.

    4.  Collect  the  information required  in  the  forms  provided,
Figure 2.4A  (English  units)  or  2.4B  (metric units), using sample
volumes equivalent  to  at  least  five  revolutions  of the dry test
meter.  Three independent runs must be made.

    5.  Calculate  Y.  for each of the three runs using Equation 2-1.
Record the values in the form (Figure 2.4A or 2.4B).
                       D  \
                        m \
+                      1 O £ I  t 4-  -L-O ^rtO^	 'M-ioO,
                      13.6/  (t, +
        Yi ='Vw  Pm + 13-6/  (td + 460WF or 273WC)       Equation 2-1

              V, P  (t  + 460°F or 273°C)
 .              d  m   w
where:

        Y. = ratio for each run of volumes measured by the wet
             test meter and dry gas meter, dimensionless
             calibration factor,

                                                 3    3
        V  = volume measured by wet test meter, m  (ft ),
         Wf

        P  = barometric pressure at the meters, mm (in.) Hg,

        D  = pressure drop across the wet tost meter, mm (in.) H20,

        t, = average temperature of dry gas meter, °C (°F),
                                                    3    3
        V, = volume measured by the dry gas meter, m  (ft  ), and

        t  = temperature of wet test meter, °C (°F).


    6.  Adjust  and recalibrate  or reject the dry gas meter if one or
more values of  Y± fall outside the interval Y _+0.02Y, where Y is the
average  for  three  runs.  Otherwise, the Y (calibration  factor)  is
acceptable and  is  to  be  used for future checks and subsequent test
runs.  The completed form should be forwarded to the  supervisor  for
approval, and then filed in the calibration log book.

    Posttest  Calibration  Check  -  After each  field  test  series,
conduct a calibration check as in Subsection 2.1.2 with the following
exceptions:

-------
Date
2-/2-Z-
                   Calibrated by
Barometer pressure, P
                                       u
Meter box number   ££  /     Wet  test meter number   /Of -/}•
                                       in. Hg    Dry gas  meter temperature correction factor
Wet test
meter
pressure
drop
.b
ft3
/. 0$&
/.on
/.0(,l
Dry test meter
gas volume
(vd),b ft3
Initial
72 5". M3
n72.
730.02!
733. /£&
Wet test
meter
gas temp
(tw),
°F
72-
72-
72-
Inlet
gas
temp

°F
76
60
80
Average
gas temp
>C oF 77 &( 32- Time of run (0),d rain ^g £6 66 Average ratio (Yt),6 /.O/J 1,017- /,o^o (Yri),f OMt, 0.110 O.W4- 3 Dm exPresse(^ as negative number. Volume passing tbrougb meter. Dry gas volume is minimum for at least five revolutions of the meter. c The average of t. and t. if using two thermometers; the actual reading if using one thermometer. H di do . The time it takes to complete the calibration run. e With Y defined as the average ratio of volumes for the wet test and the dry test meters, Y. = Y ^0.02 Y calibration and Y = Y +0.05 Y for the posttest checks; thus, for Vw (td + 460°F) [pm + (Dm/13.6)] Vd (tw (P) ?q. 1) and With Y defined as the average ratio of volumetric measurement by wet test meter to rotameter. Tolerance Y = 1 +0.05 for calibration and Y +0.1 for posttest checks. 9 (fcw + 460°F> (pm> (Eq. 3) and Yr = o, . 4) Figure 2.4A. Dry gas meter calibration data form (English units). •a O w 0) 0) 0) tQ rt- O 0)

• to


-------
Date  2 /2~2y&f    Calibrated by
        ft

Barometer pressure, Pm = 	
                                                     Meter box  number
                                        in. Hg
                                     	                    	/    Wet test meter number   /Of •


                                     Dry gas meter temperature correction factor   /1//4     °C
Wet test
meter
pressure
drop
0>m>.a
mm H~0
6.4
£.4
£.4
Rota-
meter
setting
(R8),
cc/min
4-50
4-50
450
Wet test
meter gas
volume

L
£?.75B
Z?.?gf
30.04^
Dry test meter
gas volume
(Vd),b L
Initial
/0s:fr3/
14$. $1*2-
X6/.6/?
Final
/3ST6/0
X70.377
2JI.W
Wet test
meter
gas temp
(tw),
°C
Z?-
2,2-
Z-2-
Inlet
gas
temp

Average
gas temp
(td),c
°C
Z£
68
(08
Average
ratio
(V'6
/.Ol^
/,o/e
I.OZ+
(Yri),f
AW
o.m
1.004-
  Dra exPressed as negative number.

  Volume passing through meter.  Dry gas volume  is minimum  for  at  least  five  revolutions  of the meter.

  The average of t,  and t,  if using two  thermometers;  the actual reading if using one thermometer.
                  di      do
  The time it takes to complete the calibration  run.

e'With Y defined as the average ratio of volumes  for  the wet  test  and  the  dry  test  meters,

  calibration and Y* = Y +0.05 Y for the posttest  checks;  thus,
                                                                                                 Y +0.02 Y for
     Vw (td + 273°C)
                            (Dm/13.6)]
Yl -
(t
                         (Pm)
                           (Eq. 1)
                                                    and
i. on  .     (Eq-2)
  With Y  defined as the average ratio of volumetric measurement by wet  test meter  to rotameter.


  Tolerance Yr = 1 +0.05 for calibration and Y +0.1 for posttest checks.



      v~ (t, + 273°C)  P^ + (D_/13.6) ,™J                    Y,
                                                                                                            »O D w
                                                                                                            0) 0> (D
                                                                                                            
-------
                                                   Section No. 3.15.2
                                                   Date July 1, 1986
                                                   Page 10

    1.   The leak check is not conducted  because a leak may have been
corrected that was present during testing.

    2.   Three or more revolutions of the dry gas meter may be used.

    3.   Only two independent runs need be made.

    4.   If a temperature-compensating  dry  gas  meter  was used,  the
calibration  temperature  for  the dry gas meter must be  within  6 C
(10.8 F) of the average meter temperature  observed  during the field
test series.

    When  a lower meter calibration factor is obtained as a result of
an  uncorrected  leak, the tester should correct the  leak  and  then
determine the calibration factor for the leakless system.  If the new
calibration factor changes  the  compliance status of the facility in
comparison to the lower factor,  either  include  this information in
the report  or  consult  with  the Administrator for reporting proce-
dures.   If the calibration factor does not deviate  by  >5%  from the
initial  calibration  factor Y (determined in Subsection 2.1.2), then
the dry gas meter volumes obtained during the test series are accept-
able.  If the calibration factor does deviate by >5%, recalibrate the
metering system as in Subsection 2.1.2; for the calculations, use the
calibration  factor  (initial or recalibration) that yields the lower
gas volume for each test run.

2.2  Thermometer

    The thermometer(s) on the dry gas meter inlet used to measure the
metered sample gas temperature should  be  initially  compared with a
mercury-in-glass thermometer that meets  ASTM  E-l  No.  63C  or  63F
specifications:

   1. Place  the  dial  type  or  an  equivalent  thermometer and the
mercurv-in-glass  thermometer in a hot water bath, 40 C to 50°C (104
to 122 F).  Compare the readings after the bath stabilizes.

   2.  Allow both thermometers to come to room temperature.   Compare
the readings after the thermometers stabilize.

   3.  The dial type or equivalent thermometer is acceptable  if  (1)
values agree within 3 C (5.4 F)  at both points (steps 1 and 2 above)
or (2)  if the temperature differentials at both points are within 3 C
(5.4 F) and the temperature differential is taped  to the thermometer
and recorded on the meter calibration form (Figure 2.4A or 2.4B).

   4.  Prior  to each field trip, compare the temperature reading  of
the mercury-in-glass thermometer at room temperature with that of the
thermometer that is part of  the  metering  system.  If the values or
the  corrected  values are not within 6 C  (10.8 F)  of  each  other,
replace or recalibrate the meter thermometer.

-------
                                                   Section No.  3.15.2
                                                   Date July 1, 1986
                                                   Page 11
2.3  Rotameter
   Method  7D  recommends  (optional)  that  the tester calibrate the
rotameter prior to each test.  Before  being  sent  to the field, the
rotameter  should  be  cleaned  and   maintained   according  to  the
manufacturer's  instructions.   For  this  reason,  it is recommended
(optional)  that  the calibration curve and/or rotameter markings  be
checked upon receipt  and  then  routinely  checked with the posttest
metering system check.  The rotameter may be calibrated as follows:

   1.  Ensure that the rotameter has been cleaned as specified by the
manufacturer and is not damaged.

   2.  Use the manufacturer's calibration  curve  and/or  markings on
the rotameter for the initial  calibration.   Calibrate the rotameter
as  described in the metering system calibration of Subsection 2.1.2,
and record the data on the calibration form (Figure 2.4A or 2.4B).

   3.  Use  the  rotameter  for testing  if  the  pretest  calculated
calibration is within 450 _+25 cc/min.   If,  however, the calibration
point is not  within  5%, "determine  a  new  flow  rate setting, and
recalibrate the system until the proper setting is determined.

   4.  Check the rotameter calibration  with  each  posttest metering
system check.  If the rotameter check is within 10% of the 450 cc/min
setting, the rotameter is acceptable.   If, however, the check is not
within 10% of the flow setting, disassemble and clean  the rotameter,
and perform a full recalibration.

2.4  Barometer

   The field barometer should be  adjusted  initially and before each
test series to  agree  within  2.5 mm (0.1 in.) Hg with a mercury-in-
glass barometer  or  with  the  pressure value reported from a nearby
National Weather Service Station and corrected  for  elevation.   The
tester  should  be  aware  that the pressure  readings  are  normally
corrected   to   sea  level.   The  uncorrected  readings  should  be
obtained.  The  correction  for  the elevation difference between the
weather station and the sampling point should be applied at a rate of
-2.5  mm  Hg/30  m (-0.1 in. Hg/100 ft) elevation increase,  or  vice
versa for elevation decrease.

   The calibration checks should be recorded on the pretest  sampling
form (Figure 2.5).

2.5  Analytical Balance

   The analytical balance used  to weigh the reagents for the nitrate
stock standard should be calibrated by the following procedure:

   1.  Zero the balance.

-------
                                               Section No. 3.15.2
                                               Date July 1, 1986
                                               Page 12
Date     Z-Z^-	   Calibrated by
         '    /                            "
Meter box number £.£.' — /


Dry Gas Meter*

Pretest calibration factor =  /.&ZQ  (within 2% of average factor
  for each calibration run).

Rotameter

Pretest calibration factor (Y ) or setting =   /,OQ    (between
  400 and 500 cc/min).

Dry Gas Meter Thermometer

Was a pretest temperature correction made? 	yes  X  no

If yes, temperature correction 	 (within 3°C (5.4°F) of
  reference values for calibration and within 6 C (10.8 F) of
  reference values for calibration check).

Barometer

Was the pretest field barometer reading correct?  X  yes 	no
  (within 2.5 mm (0.1 in.) Hg of mercury-in-glass barometer).
*Most significant items/parameters to be checked.
              Figure 2.5.  Pretest sampling checks.

-------
                                                   Section No. 3.15.2
                                                   Date July 1, 1986
                                                   Page 13

    2.  Place  a  5-g  Class-S  weight  on  the  balance.  Record the
        balance reading for the 5-g weight.

    3.  Place  a  10-g  Class-S  weight  on the balance.  Record  the
        balance reading for the 10-g weight.

    4.  The balance readings for the 5-g  and  10-g  weights  must be
        within 2 mg of the actual weights.

    5.  If the balance readings are greater than +2 mg either  of the
        actual weights, repair the  balance  or  contact  the balance
        manufacturer.

2.6  Ion Chromatograph System

2.6.1  Performance Check of the Ion  Chromatograph - Method 7D states
that  the  instrument  used  for  analysis  should  provide  adequate
resolution  of  NOo" and should be able to resolve and detect nitrite
ion (N02 ).  It  is  recommended  that  the instrument be performance
checked prior to initial  use  to ensure that the instrument can meet
the above criteria.

    Method  7D does not quantify the criteria for acceptable  instru-
ment performance.  The numerical limits  and  procedures  given below
are  offered  from  a purely technical viewpoint.   Their  observance
should  ensure that the instrument  conforms  with  the  method,  but
should  not  be  interpreted  as  a requirement.  Considerations  for
preparing the ion Chromatograph for analysis follow.

    Conductivity Detector  -  Prior  to  its  initial  use,  the con-
ductivity detector of the ion Chromatograph (1C) should be calibrated
by  the  method  described in the operator's manual.  Following  this
calibration  it  is  highly  recommended that the analyst  conduct  a
preliminary  calibration  of  the 1C.  Because the 1C calibration  is
conducted concurrently with the field sample analysis when performing
Method  7D,  the  full  discussion  of the calibration  procedure  is
presented in Section 3.15.5.

    Recorder  -  A  strip chart recorder compatible with  the  output
voltage range of the conductivity detector  may be used to record the
ion chromatogram.  Manual measurement techniques that can be used for
quantitation of  the  chromatogram  include (a) peak height, (b) peak
area by triangulation, (c) peak area by multiplying peak height times
the peak width at half-height, (d) peak  area by cutting out the peak
from the chromatogram and weighing it on an analytical  balance,  and
(e) peak area by planimetry.

    The use of an electronic integrator, if available, is recommended
for greater accuracy and precision.  The electronic integrator can be
used in the peak area mode when the integration parameters are set up
properly.   The key integration parameters for peak area determination
concern the identification of the beginning and end of a peak and the

-------
                                                   Section No. 3.15.2'
                                                   Date July 1, 1986
                                                   Page 14

placement of the baseline under the peak.  Analysts  should carefully
read the operator's manual and understand the selection and set up of
the  integration  parameters  for  their  particular integrator.  The
electronic  integrator  can  also  be  used in the peak  height  mode
provided that the peaks are symmetrical  and  an  acceptable standard
calibration curve can be generated.

    Sample  Injection  Device  Contamination  Check - The analyst  is
encouraged to check the  sample injection device for contamination by
injecting  water  before  the  calibration  standards  are  analyzed.
Contaminants  will  appear  as  peaks  on the chromatogram.  Repeated
injections of water should be used to remove  contaminants  from  the
sample  injection  device.   If  certain peaks remain  after  several
injections,  then  the  water  may  be  contaminated  and  should  be
replaced.

    Separation  of  Nitrate, NO^  - To ensure accurate  results  from
the  ion  chromatographic   analysis, baseline separation of the N03~
peak  from the other ion peaks should be achieved.  A source of  SO."
in a sample may be sulfur dioxide  present  in  the  effluent  stream
sample.  Figures  2.6a  and_ 2.6b  show two chromatograms, one having
overlapping N0«   and   SO."  peaks,  and  the  other having baseline
separation of the  N03  ana SO." peaks.

    The  analyst is encouraged to check the performance  of  the  ion
chromatograph system before analyzing samples in order to ensure that
baseline separation of NO3  is attainable.

    The ion chromatograph can be performance checked using a solution
containing  NO~   and_ SO.   for_ compliance  purposes  or a solution
containing N03 ,  N0o~'  ^nc* SO." if the nitrate is to be quantified.
A solution  that will provide rigorous conditions involves the use of.
KNO~ working standard  solution £described in Section 3.15.4, page 3)
and NO2~ (if applicable) and SO." solutions, the preparation of which
are addressed below.

    The S0.~ solution is prepared as  follows:  Weigh  out  0.231  of
sodium suitate (Na2SO.), and transfer it  to  a beaker.  Dissolve the
Na2SO.  in  water,  quantitatively  transfer the solution to a 250-ml
volumetric flask, and finally, dilute to the mark with water.

    The  concentration  of  the solution is 625 vg  SO."/ml.   Sodium
sulfate (Na2SO.) is a component of  the  pusher solution in the Orsat
apparatus used for Method 3.  It is  not  special and has been chosen
because  of  its probable availability.  Other SO." reagents  can  be
used.

    If  the  nitrite is to be  quantified,  then  separation  of  the
nitrate  peak  should  also  be  checked.  To  prepare the N02  stock
solution,  first  weigh  out  52.5  mg  sodium  nitrite  (NaN02)  and

-------
                                              Section No.  3.15.2
                                              Date July 1,  1986
                                              Page 15
                                                          so,
WATER DIP
   Figure 2.6a.  Example chromatogram having overlapping peaks.
   Figure 2.6b.  Example chromatogram showing baseline
                 separation of peaks.

-------
                                                   Section No. 3.15.2
                                                   Date July 1, 1986
                                                   Page 16

transfer it to a beaker.  Dissolve the NaN02 in water, quantitatively
transfer  it to a 250-ml volumetric flask, and finally, dilute to the
mark with water.  To prepare the N02~ working solution, pipet 10.0 ml
of the stock solution into a 100-ml  volumetric  flask, and dilute to
volume_with water.  The concentration of the working  solution  is 14
yg N02~/ml.

   To prepare the performance check solution, pipet 10 ml of the KN03
working standard solution, 8 ml of the SO." solution, and 1 ml of thS
N02~ working solution (if applicable) into a 200-ml volumetric flask,
ana dilute to the mark with water.

   The_concentration  of  N03~ in the performance check sample is 7.5
yg NO3~/ml,  which  corresponds  to  a  NO  level around the emission
standard for coal-fired boilers subject to 40 CFR Part 60, Subparts D
or  Da.   This  correspondence  also is based on the assumptions that
sampling  is  conducted  for  one  hour at 500 ml/min  and  that  the
effluent sample is 12% (v/v) C02.

   The sulfate concentration of the performance check sample is 25 yg
SO. /ml, which corresponds to an S02 level of roughly 1000 ppm (for a
1-nour sample acquired at 500 ml/min and containing 12%  (v/v)  C02).
This concentration level should be more than adequate  for situations
involving  the  application of Method 7D to sources subject to 40 CFR
Part  60, Subpart D; thus, it is recommended that  analysts  decrease
the SO." concentration in proportion  to  the S02 levels expected for
the effluent.  For example, if the effluent concentration of S02 were
500 ppm, 5 ml (rather than 10 ml) of the SO." solution would  be used
in preparing the performance check sample.  For applications upstream
of flue gas desulfurization systems at sources subject to 40 CFR Part
60,  Subpart  Da,  the  opposite  situation  may  exist,  and  it  is
recommended that the concentration of SO." be increased accordingly.

   The_N02~ concentration of the performance  check  solution is 0.07
yg N02~/ml.   This  corresponds  to  6  ppm N02 for a one-hour sample
acquired at 500 ml/min and containing 12% (v/v; C02.

   The  performance  check  solution  should  be  analyzed  with  the
calibration   standards   during   the   initial  check  of  the  ion
chromatograph's  calibrations.    The  same  experimental  conditions
should be observed for the solution and the standards.  Figure 2.7 is
an example chromatogram showing where the N02 , N03 , and SO." can be
expected to elute.

2.6.2   Preparation of Calibration Curve - Method  7D  gives  general
instructions  for  preparing  the  calibration  curve  for   the  ion
chromatograph.  Accordingly, the method requires that:

   (a) at least four calibration standards be prepared;

-------
                                               Section No. 3.15.2
                                               Date July 1, 1986
                                               Page 17
                            5.5 minutes
                     N03   3.7 minutes
                                                 NO_  1.4 minutes
            Inject
 Figure 2.7.   Chromatogram showing resolution of nitrite,
              nitrate,  and sulfate peaks.

(b) the concentration range of the calibration standards cover the
    concentration range of the samples being analyzed;

(c) the  calibration  standards be prepared from the KN03 standard
    solution using pipettes having volumes 1.0 ml or greater;

(d) the calibration  standards  be  analyzed  and  the  results be
    interpreted  in  the  same  manner as for  the  samples  being
    analyzed;

(e) the results of the analyses of  the  calibration standards (in
    units of either  peak  height  or peak area) should be plotted
    versus the standards' concentrations (in units of yg N03~/ml);

(f) the plotted points define a linear relation;

(g) the calibration  equation  be determined from the points using
    linear regression;  and

-------
                                                   Section No.  3.15.2
                                                   Date July 1,  1986
                                                   Page 18

    (g)  the calibration equation be determined from the points using
         linear regression;  and

    (h)  the calibration standards  be  analyzed  twice  in  order to
         compensate  for  any  drift  in  the  response  of  the  ion
         chromatograph.

    The method leaves to the analyst details including:

    (a)  the concentration  values  for  the  individual  calibration
         standards;

    (b)  the degree of linearity of the calibration  curve  that will
         ensure quality results; and

    (c)  the procedure to  be  used to compensate results for the ion
         chromatograph's drift.

    Concentration values for calibration  standards  -   The  step-by
step-procedures for preparing the calibration standards and preparing
the calibration curve are given in Section 3.15.5.

-------
                                                           Section No. 3-15.2
                                                           Date July 1. 1986
                                                           Page 19
            Table 2.1.  ACTIVITIY MATRIX FOR CALIBRATION OF EQUIPMENT
Apparatus
 Acceptance Limits
Frequency and method
    of measurement
Action if
requirements
are not met
Wet test meter
Capacity of at least 2
L/min and an accuracy
within l.Q%
Calibrate initially and
then yearly by liquid
displacement
Adjust until
specifications
are met, or
return to
manufacturer
Dry gas meter
Y  = 'Y+0.02Y at a
flow rate of about
450 cc/min
Calibrate vs. wet test
meter initially and
when the posttest check
is not within Y+0.05
Repair and
then recali-
brate, or
replace
Dry gas meter
 thermometer
Within 3°C (5-t F)
of true value
Calibrate each initially
as a separate component
against a mercury-in-
glass thermometer; after
train is assembled
before each field test,
compare with mercury-in-
glass thermometer
Adjust, deter-
mine a con-
stant correc-
tion factor
or reject
Rotameter
Clean and maintain ac-
cording to manufactur-
er's instructions (re-
quired) ; calibrate to
+5# (recommended)
Initially and after each
field trip
Adjust and re-
calibrate, or
reject
Barometer
Within 2.5 mm
(0.1 in.) Hg of
mercury-in-glass bar-
ometer or of weather
station value
Calibrate initially
using a mercury-in-glass
barometer; check before
and after each field
test
Adjust to
agree with
certified
barometer
Analytical
  balance
Weight within 2 mg of
standard weights
(Class-S)
Use standard weight be-
fore preparation of
working solution
Repair or
return to
manufacturer
 (continued)

-------
                                                           Section No. 3.15.2
                                                           Date July 1, 1986
                                                           Page 20
Table 2.1.  (continued)
Apparatus
Acceptance Limits
Frequency and method
    of measurement
Action if
requirements
are not met
Ion chromato-
  graph
Calibrate prior to each
set of sample analyses
With each set of field
samples;  calibration
standards prepared from
potassium nitrate
Interpret data
using another
technique;
e.g., if using
peak height,
change to peak
area; analyze
additional
calibration
standards;
calibrate
conductivity
detector;
consult opera-
tor's manual

-------
                                                   Section No.  3.15.3
                                                   Date July 1,  1986
                                                   Page 1
3.0  PRESAMPLING OPERATIONS
    The quality  assurance activities for presampling preparation are
summarized in  Table 3.1 at the end of this section.  See Section 3.0
of this Handbook for details on preliminary site visits.

3.1  Apparatus Check and Calibration

    Figure 3.1 or  a similar form is recommended to aid the tester in
preparing  an  equipment  checklist,  status report form, and packing
list.

3.1.1  Sampling Train - The schematic for the NO   sampling  train is
given  in  Figure  1.2.   Commercial models of tftis system are avail-
able.  Each individual or fabricated train must be in compliance with
the specifications in Method 7D, Section 3.15.10.

3.1.2   Probe  -  The  probe should be cleaned internally by brushing
first  with  tap  water,  then  with water, and finally with acetone.
Allow the probe to dry in  the  air.   In extreme cases, the glass or
stainless steel  liner  can  be  cleaned  with stronger reagents; the
objective is to leave the liner free from contaminants.  The  probe's
heating system should be checked  to  see  whether  it  is  operating
properly.  The probe  must  be  leak free when sealed at the inlet or
tip and checked for leaks at a vacuum of 250 mm (10 in.) Hg with  the
meter box.   Any  leaks  should  be  corrected.   The liner should be
sealed inside  the  metal sheath to prevent diluent air from entering
the source since most stacks are under negative pressure.

3.1.3   Restricted Orifice  Impingers  and  Glass  Connectors  -  All
glassware should be cleaned with  detergent  and  tap water, and then
with  water.  Any items that do not  pass  a  visual  inspection  for
cracks or breakage must be repaired or discarded.

3.1.4  Drying Tubes - Drying  tubes  should  be  packed  with  6-  to
16-mesh silica gel and sealed at both ends.

3.1.5  Valve and Rotameter  - Prior to each field trip or at any sign
of  erratic behavior, the flow control valve and rotameter should  be
cleaned  according  to the maintenance procedure recommended  by  the
manufacturer.

3.1.6   Pump  - The vacuum pump  and  oiler  should  be  serviced  as
recommended by the manufacturer,  every  3 months, or every 10th test
(whichever  comes  first),  or upon erratic behavior  (nonuniform  or
insufficient pumping action).

3.1.7  Dry  Gas  Meter  - A dry gas meter calibration check should be
made  in  accordance  with  the  procedure  in  Section  3.15.2.   An
acceptable posttest check from the previous test is sufficient.

-------
                                               Section No. 3.15.3
                                               Date July 1, 1986
                                               Page 2
Apparatus check
Probe
Type liner
Glass *X
Stainless
steel
Other
Heated properly*
Leak checked
Filter
Glass wool
Other

Glassware
Restriced
orifice
impinger
Size
Type

Meter System
Leak- free pumps*
Rate meter*
Dry gas meter*
C00 Measurement
OrSat t/"
Fyrite

Reagents
Water
Potassium
permanganate*
Silica gel
Other
Barometer
Drying tube
Acceptable
Yes
iX
tX"
ix^
^
i/
iX
X'
^
i/
i/
«X"
X
^
No







Quantity
required
+
^^i^ll tooyt.
14-
•2.
/
1
2^
2 j*l
2- J*l
Z&
/
6>
Ready
Yes
l^
^
^
(^
iX
\^
i^
^
i/'
^
(^
t^
NO







Loaded
and packed
Yes
^
^
^
iX
i^
^
^
^
^
^
No







* Most significant items/parameters to be checked.
                Figure 3.1.  Pretest preparations.

-------
                                                   Section No.  3.15.3
                                                   Date July 1, 1986
                                                   Page 3

3.1.8  Thermometers - The thermometers should be  compared  with  the
mercury-in-glass thermometer  at room temperature prior to each field
trip.

3.1.9  Barometer - The  field  barometer  should be compared with the
mercury-in-glass barometer or with a National Weather Service Station
reading prior to each field trip.

3.1.10   COg Analysis - Method 3 sampling apparatus  should  be  leak
checked, ana the reagents should be checked  to ensure freshness (see
Section 3.2 of this Handbook).

3.2  Reagents for Sampling

    The following reagents are needed during the  sampling  phase  of
Method 7D:

3.2.1   Water - Deionized distilled  water  should  conform  to  ASTM
specification D1193-82, Type III (see Subsection  1.4.1  for detailed
specifications).

3.2.2 Potassium Permanganate/Sodium Hydroxide (KMnO./NaOH) Solution -
Dissolve 40.0 g of KMN04 and 20.0 g of NaOH in 940 ffll of water.

3.3  Packaging Equipment for Shipment

    Equipment  should  be packed in rigid containers  to  protect  it
against rough handling during shipping and field operations.

3.3.1 Probe - The inlet and  outlet  of  the probe must be sealed and
protected from breakage.   A  suggested  container  is  a wooden case
lined with polyethylene foam or other  suitable packing material; the
case should have separate compartments  for  individual devices.  The
case should be equipped with handles or eye hooks  that can withstand
hoisting, and should  be  rigid to prevent bending or twisting during
shipping and handling.

3.3.2 Impingers, Connectors,  and  Assorted Glassware - All impingers
and glassware  should be packed in a rigid container and protected by
polyethylene  foam  or  other  suitable packing material.  Individual
compartments for glassware help to organize  and  protect  each item.
The impinger train may be charged and assembled in the laboratory  if
sampling is to be performed within 24 hours.

3.3.3 Drying Tubes and Volumetric Glassware - A rigid container lined
with  polyethylene foam material protects drying tubes  and  assorted
volumetric glassware.

3.3.4 Meter Box - The meter box, which contains the valve, rotameter,
vacuum pump, dry gas meter, and thermometer(s), should be packed in a
rigid shipping container  unless  its  housing  is  strong  enough to

-------
                                                   Section No.  3.15.3
                                                   Date July 1, 1986
                                                   Page 4

protect  components  during travel.  Additional pump  oil  should  be
packed if oil is required  for  operation.  It is advisable to ship a
spare meter box in case of equipment failure.

3.3.5 Wash Bottles and  Storage  Containers  - Storage containers and
miscellaneous  glassware should  be  packed  in  a  rigid  foam-lined
container.   Samples  being transported in the containers  should  be
protected from extremely low ambient temperatures (below freezing).

-------
                                                                Section No. 3.15-3
                                                                Date July 1, 1986
                                                                Page 5
               Table 3.1.  ACTIVITY MATRIX FOR PRESAMPLING OPERATIONS
Operation
 Acceptance limits
Frequency and method
    of measurement
Action if
requirements
are not met
Probe
1. Probe liner free
of contaminants
                  2. Probe leak free at
                  250 inm (10 in.) Hg

                  3. No moisture conden-
                  sation
1. Clean probe inter-
nally by brushing with
tap water, then deion-
ized distilled water,
then acetone; allow to
dry in air before test

2. Visually check for
cracks before test

3. Check out heating
system initially and
when moisture appears
during testing
1.   Retrace
cleaning pro-
cedure and
assembly
                                                  2.  Replace
                                                  3.  Repair
                                                  or replace
Restricted
 orifice impin-
 gers and glass
 connectors
Clean and free of
breaks, cracks, etc.
Clean with detergent,
tap water, and then
with deionized dis-
tilled water
Repair or
discard
Flow control
 valve and
 rotameter
Clean and without sign
of erratic behavior
(ball not moving)
Clean prior to each
field trip or upon
erratic behavior
Repair or
return to
manufacturer
Vacuum pump
Maintain sampling rate
of 400 to 500 cc/min
at a vacuum up to
250 mm (10 in.) Hg
Service every 3 mo. or
upon erratic behavior;
check oiler jars every
10th test
As above
Dry gas meter
Clean and within 2%
of calibration factor
Calibrate according
to Section 3.15.2;
check for excess oil
if oiler is used
As above
CO- analyzer
Leak-free and fresh
reagents
Leak check, and check
reagents
As above
(continued)

-------
                                                              Section No. 3-15-3
                                                              Date July 1, 1986
                                                              Page 6
Table 3.1  (continued)
Operation
 Acceptance limits
 Frequency and method
     of measurement
Action if
requirements
are not met
Reagents

Sampling
Requires all ACS grade
reagents
Prepare and store in
sealed containers
Prepare new
reagent
Sample recovery
Requires water on
site
Quantity sufficient to re-
cover sample after testing
and clean impingers prior
to testing
Prepare new
reagent
Package Equip-
ment for Ship-
ment

Probe
Protect with poly-
ethylene foam
Prior to each shipment
Repack
Impingers,
 connectors,
 and assorted
 glassware
Pack in rigid con-
tainers with poly-
ethylene foam
As above
As above
Drying tubes,
 volumetric
 glassware
Sturdy container
lined with foam
As above
As above
Meter box
Meter box case and/or
container to protect
components, pack spare
meter box and oil
As above
As above
Wash bottles
 and storage
 containers
Pack in rigid foam-
lined container
As above
As above
Samples
Protect from extreme
cold (below freezing)
As above
As above

-------
                                                   Section No.  3.15.4
                                                   Date July 1, 1986
                                                   Page 1
4.0  ON-SITE MEASUREMENTS
    On-site activities include transporting the equipment to the test
site, unpacking  and  assembling,  sampling  for nitrogen oxides, and
recording  the data.  The quality assurance activities are summarized
in Table 4.1 at the end of this section.

4.1  Transport of Equipment to the Sampling Site

    The  most  efficient  means  of transporting the  equipment  from
ground level to the sampling site  (often  above ground level) should
be decided during the preliminary site  visit  or  by  prior  corres-
pondence.  Care should be taken to prevent damage to the equipment or
injury to test personnel during the moving.  A laboratory area should
be  designated  for  preparing the absorbing reagents,  charging  the
impingers, and sample recovery.

4.2  Preliminary Measurements and Setup

    Method 7D outlines the procedure  for  determining the concentra-
tion of nitrogen  oxides  in  the  gas  stream.   The accuracy of the
equipment that has been transported to the sampling site and that may
have been handled roughly can be determined  by  making  a  one-point
check of the rotameter reading against the dry gas meter  reading  at
the test site.  Use Equation 3 in Figure 2.4A or 2.4B, and substitute
dry gas meter readings in place of wet test meter readings (i.e.,  V,
= V ).  Y .   should be between 0.9 and 1.1; if not, the meter box hai
losr its rate or volume calibration.   The  tester  can still use the
meter  box,   but  the data should not be released for decision making
until a post-test recalibration  has been made.  If the dry gas meter
calibration  factor did change, the dry gas meter volumes may have to
be corrected.  Record the test identification  number on the sampling
data form, Figure 4.1.

4.3  Sampling

    The on-site sampling includes the following steps:

    1. Preparation and/or addition of  the  absorbing reagents to the
impingers.

    2.  Setup of the sampling train.

    3.  Connection to the electrical service.

    4.  Preparation of the probe (leak check of entire sampling train
and addition of particulate filter).

    5.  Check of rotameter setting.

    6.  Insertion of the probe into the stack.

-------
Plant name
Location
Operator
                        ro
                         oyer'
                         /Y&.
                         City
                         Date
                                                             Section No.  3.15-4
                                                             Date July 1,  1986
                                                             Page 2
                                            Sample no.
                                                           A?-/
Probe length/material 	
Meter box no.     c^5 —/
                                            Probe setting 	
                                            Meter factor (Y)
Sampling point location(s)
Rotameter setting
                                 nv*
                         Bar press  mm  (in.)  Hg
                         Rotameter  check?
                                                                    2-tf. 4-1
Initial leak check?
CO- concentration
(1)    Af. g   (2)
                                            Final leak check? 
IbWfa
l(t7.ZZ(,


Total
non
Sample flow
rate setting,
cc/min (ft^/min)
	
4Zb
4$o
4Q
4^0
4^0
4$0
4- so
t&>
4ZO
4-^0
4^o
4*^0



Sample volume
metered,,j(V )
L (ft3) m
— •
2. 14J*
2-.14V
2.Z&-
2-ZW
2-ZZ1-
2,2^
2-Z^-~
Z.Z^°I
Z.Ztt
Z^W>
2-^4
Z.Z.44-


V
m
avgZ.Z^7
Percent
deviation, a
%
	
-O.Z2.
-0.41
+O.Z7
-0.31
-H?-04-
-0.^1
+0.13
-+0.&0
+o.?>\
-0,13
to.l3>
-O.3/


Avg
dev "0. 42
Dry gas
meter temp,
§C <°F)
—
72
74
74
75-
7b
76
76
7g
70
7?
76
7?


Avg
76-4-
Percent deviation =   m  "   m avg x 100  (must be less than 10 percent).
                         V  avg
                   Figure 4.1. Field sampling data form for NO .

-------
                                                   Section No.  3.15.4
                                                   Date July 1,  1986
                                                   Page 3

    7.  Sealing of the port.

    8.  Check of the temperature of the probe.

    9.  Sampling.

   10.  Measuring the C02 concentration.

   11.  Recording of the data in Figure 4.1.

A final leak check of the train is always performed after sampling.

4.3.1   Preparation  and/or  Addition   of   Absorbing   Reagents  to
Collection System - Absorbing  reagents  can  be prepared on site, if
necessary, according to the directions in Subsection 1.4.1.

    1.  Use a pipette or a graduated cylinder  to introduce 200 ml of
alkaline permanganate (KMnO./NaOH)  solution  into  each of the three
impingers.

    2.  Place in  the  sampling  train  a drying tube that has new or
regenerated silica gel.

4.3.2  Assembling the Sampling Train - After assembling  the sampling
train as shown in Figure 1.2, perform the following:

    1.  Adjust probe heater to operating  temperature.  Place crushed
ice and water around the impingers.

    2.   Leak check the sampling train  just  prior  to  use  at  the
sampling  site   (not  mandatory) by temporarily attaching a rotameter
(capacity of 0 to 40 cc/min) to the outlet of the dry  gas  meter and
placing a vacuum gauge  at  or  near the probe inlet.  Plug the probe
inlet,  pull  a  vacuum  of at least 250 mm (10 in.) Hg, and note the
flow  rate indicated by the rotameter.  A leakage rate £  2%  of  the
average  sampling rate is acceptable.  Note;  Carefully  release  the
probe inlet plug before turning off the pump.  It  is  suggested  (but
not mandatory) that the pump be leak checked separately, either prior
to or after the sampling run.  If prior to the  run,  the  pump   leak
check shall precede the  leak check of the sampling train.  If after,
the pump leak check shall follow the train leak check.  To leak check
the pump, proceed  as  follows.   Disconnect the drying tube from the
probe impinger assembly.  Pull a vacuum of 250 mm (10 in.) Hg.    Plug
or pinch off the outlet  of  the  flow  meter,   and then turn off the
pump.  The vacuum should remain stable for at least 30 seconds.

    3.  Place a  loosely packed filter of glass wool in the end of the
probe, and connect the probe to the first impinger.

4.3.3  Rotameter  Setting Check (Optional) - After leak checking  the
sampling train, disconnect the  probe  from  the  first impinger, and
connect  the  filter (optional).  The filter  is  a  tube  containing

-------
                                                   Section No. 3.15.4
                                                   Date July 1, 1986
                                                   Page 4

approximately 20 g of 5-Angstrom  molecular  sieve  to remove the NO
from  the  ambient air.  Start the pump, and adjust the flow  to  the
rotameter setting to be used during the sampling run.  After the flow
has stablized, start measuring the volume sampled, as recorded by the
dry  gas  meter and the sampling time.  Collect sufficient volume  to
measure  accurately the flow rate, and calculate the flow rate.   The
average flow rate must be less  than  500 cc/min for the sample to be
valid; therefore, it is recommended that the flow rate be checked  as
above prior to each run.  Record the sampling rate on the data form.

4.3.4 Sampling (Constant Rate) - Sampling is performed  at a constant
rate  of  between  400  and 500 cc/min as indicated by the  rotameter
during the entire sampling run.  The procedure is as follows:

    1.  Record the initial dry gas meter readings, barometer reading,
and other data as indicated in Figure 4.1.  Double  check the dry gas
meter reading.

    2.  Position the tip of the probe at the sampling  point, connect
the probe  to  the  first impinger, and start the pump.  Warning;  If
the stack is under a negative  pressure of >250 mm (10 in.) H20 while
disconnected from the impinger, the probe should be positioned at the
sampling  point,  the  sample  pump  turned  on,  and  then the probe
immediately  connected  to  the  impinger  to  prevent  the  impinger
solutions from being siphoned backwards.

    3.  Adjust the sample flow to the  preselected  flow rate (400 to
500 cc/min) as indicated by the rotameter.

    4.   Maintain  a  constant  rate  within  10%  during  the entire
sampling run, and take readings  (dry  gas meter, temperatures at dry
gas meter, and rate meter) at least every 5 minutes.

    5.  Refer to emission standards  for minimum sampling time and/or
volume.   (For  example,  the  Federal standard for fossil fuel-fired
steam generators specifies a minimum sampling time of 60 minutes; for
relative accuracy tests,  when  the S02 concentration is greater than
1200  ppm, the sampling time should be 30 minutes since S0>2  depletes
the  absorbing solution.)  A quick calculation should be  made  after
half the sampling time to guarantee  that  the sampling rate will not
exceed 500 cc/min.

    6.  During sampling,  measure  the  C02  content of the stack gas
near the  sampling  point  using  Method  3.   The  single-point grab
sampling procedure is adequate, provided the measurements are made at
least three times (near the start, midway, and before the  end  of  a
run)  and  provided the average C02 concentration  is  computed.   An
Orsat (which is highly  recommended)  or  Fyrite analyzer may be used
for this analysis.   The  results should be recorded on the data form
(Figure 4.1).

-------
                                                   Section No. 3.15.4
                                                   Date July 1, 1986
                                                   Page 5

    7.  Turn off the pump at the conclusion  of  each run, remove the
probe from the stack, and record the final readings.  Warning; Again,
if  the stack is under negative pressure, disconnect the probe first,
and turn off the pump immediately thereafter.
    8.    Conduct a leak
( mandatory ) .
                         check,  as  described  in  Subsection  4.3.2
    9.   Calculate  the  sampling rate.  The sample volume ( V )  for
each point should be within  10%  of  the average sampling volume for
all points, and the average sampling rate for the test should be less
than 500 cc/min.  If the average sampling  rate  exceeds  500 cc/min,
the sample collection efficiency may be affected.

4.4  Sample Recovery

    Method 7D requires  transfer  of  the  impinger  contents and the
connector washings to a  polyethylene storage container.  This trans-
fer should  be done in the "laboratory" area to prevent contamination
of the test sample.

    After  completing the final leak check, disconnect the impingers,
and transport them to the cleanup area.  Cap off the impinger section
with  the use of polyethylene or equivalent caps before transport  to
the  cleanup  area.   Transfer  the contents of the impingers into   a
labeled,  leak-free  polyethylene  sample  bottle.  Rinse  the  three
impingers a couple of times and the connecting  tubes once with 3- to
15-ml  portions  of water.  Add these washings  to  the  same  sample
bottle, and mark the fluid  level on the side.  Place about 100 ml of
the  absorbing  reagent  (KMn04/NaOH) in a polyethylene  bottle,  and
label it for use  as  a  blank  during samp?.e analysis (once for each
test).  An example of a sample label is shown in Figure 4.2.
Plant City
Site Sample type
Date Run number
Front rinse [J Front filter LJ Front solution LJ
Back rinse LJ Back filter LJ Back solution 1 	 ]
• •
Solution Level marked J2
j_i
Volume: Initial Final 2
Cleanup by £





























              Figure 4.2.  Example of a sample label.

-------
                                                   Section No. 3.15.4
                                                   Date July 1, 1986
                                                   Page 6
4.5  Sample Logistics (Data) and Packing Equipment
    The sampling and sample  recovery  procedures  are followed until
the required number of runs are  completed.   Log  all  data  on  the
Sample  Recovery  and  Integrity  Data  Form,  Figure  4.3.   If  the
impingers,  and  connectors  are  to  be  used in the next test, they
should be rinsed with water, and a new drying tube should be inserted
into the sampling train.  At the completion of the test:

    ~1.  Check all sample  containers for proper labeling (time, date,
location,  number  of  test,  liquid  level,  and any other pertinent
documentation). Be sure that a blank has been taken.

    2.  Record all data collected during the field test in  duplicate
by using carbon paper or by using data forms and a  field  laboratory
notebook.  One set of data should be mailed to the  base  laboratory,
given  to another team member, or given to the Agency.  Hand carrying
the  other  set  (not  mandatory)  can  prevent  a  very  costly  and
embarrassing mistake.

    3.   Examine  all  sample  containers  and sampling equipment for
damage, and pack them for shipment  to  the  base  laboratory,  being
careful  to  label all shipping containers to prevent loss of samples
or equipment.

    4.   Make  a  quick  check  of  the  sampling and sample recovery
procedures using the data form,  Figure 4.4.

-------
                                               Section No.  3.15.4
                                               Date July 1,  1986
                                               Page 7
Plant
              ro der   /A**/-     Sampling location
                                                          A/0 . 3
                        Field Data Checks
Sample recovery personnel
Person with direct responsibility for recovered samples
Sample
number
1
2
3
Blank
Sample
identification
number
M-l



Date
of
recovery
3/75/04



Liquid
level
marked
i^



Stored
in locked
container
^^



Remarks
                               fw
Signature of field sample trustee
                      Laboratory Data Checks
Lab person with direct responsibility for recovered samples
Date recovered samples received
Analyst     ^coft
Sample
number
1
2
3
Blank
Sample
identification
number
AP-I



Date
of
analysis
3//B/84
/ i


Liquid
level
marked
^-



Stored
in locked
container
t_^



Remarks
                                  n
Signature of lab sample trustee
         Figure 4.3.  Sample recovery and

-------
                                               Section No. 3.15.4
                                               Date July 1, 1986
                                               Page 8
Sampling

Impinger contents properly selected, measured, and placed in
impingers?* 	^	

Impinger Contents/Parameters*

1st: 200 ml of KMn04/NaOH 	
2nd: 200 ml of KMn04/NaOH 	

3rd: 200 ml of KMn04/NaOH 	r
Drying tube: 6- to 16-mesh silica gel 	^_

Probe heat at proper level?* 	_

Pretest leak check at 250 mm (10 in.) Hg?

Leakage rate? 	0- OO4-
Check of rotameter setting?
Probe placed at proper sampling point?
Flow rate constant at approximately 450 cc/min?*

CO,, concentration measured?* 	  £-
Sample Recovery

Contents of impingers placed in polyethylene bottles?

Fluid level marked?*                       """^
Sample containers sealed and identified?*
* Most significant items/parameters to be checked.
                  Figure 4.4.  On-site measurements.

-------
                                                              Section No.  3.15-4
                                                              Date July 1,  1986
                                                              Page 9
            Table 4.1.  ACTIVITY MATRIX FOR ON-SITE MEASUREMENT CHECKS
Activity
   Acceptance limits
Frequency and method
    of measurement
Action if
requirements
are not met
Preparing and/
or adding
absorbing
reagents
Add 200 ml of
NaOH to the impingers
Add 40.0 g of KMnO. and
20.0 g of NaOH to §40 ml
of water
Reassemble
collection
system
Assembling
the sampling
train
1.  Assemble to speci-
fications in Fig. 1.2

2.  A leakage rate
of <2% of the average
sampling rate
1.   Before each sampling
run

2.   Leak check before
sampling (recommended) by
attaching a rotameter to
dry gas meter outlet,
placing a vacuum gauge
at or near probe inlet,
and pulling a vacuum
of > 250 mm (10 in.) Hg
1.  Reassemble
                                                                      2. Correct
                                                                      the leak
Sampling (con-
 stant rate)
1.  Within 10% of
constant rate
                2.  Minimum accepta-
                ble time is 60 min
                and sampling rate
                less than 500 cc/min

                3.  Less than 2% leak-
                age rate at 250 mm
                (10 in.) Hg
                4.  Determine CO,
                content
1.  Calculate % devia-
tion for each sample
using equation in
Fig. 4.1
                          2.  Make a quick calcu-
                          lation prior to comple-
                          tion and an exact calcu-
                          lation after completion

                          3.  Leak check after
                          sampling run (mandatory);
                          use same procedure
                          as above
                          4.  Measure C0_ content
                          using Method 3
1.   Repeat
the sampling,
or obtain ac-
ceptance from
a representa-
tive of the
Administrator

2.   As above
                                As above
                                As above
(continued)

-------
                                                              Section No. 3.15.4
                                                              Date July 1, 1986
                                                              Page 10
Table 4.1  (continued)
Activity
Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Sample logistics
  (data) and
  packing of
  equipment
1.  All data are re-
corded correctly
                  2.  All equipment ex-
                  amined for damage and
                  labeled for shipment
                  3.  All sample con-
                  tainers properly
                  labeled and packaged
1.  Visually check upon
completion of each run
and before packing
                        2.  As above
                        3.  Visually check upon
                        completion of test
1.  Complete
the data form
                          2.  Redo test
                          if damage
                          occurred
                          during testing

                          3.  Correct
                          when possible

-------
                                                   Section No. 3.15.5
                                                   Date July 1, 1986
                                                   Page 1
5.0  POSTSAMPLING OPERATIONS
    The postsampling operations  for  Method  7D include an apparatus
check, a barometer check, sample preparation,  and sample analysis by
ion  chromatogr aphy .   The procedures for the apparatus check and the
barometer check  are  the  same as in Method 6.  These procedures are
detailed in Method  6  in  Section  3.5.5 and are not discussed here.
The  procedures  for  sample  preparation  and  sample  analysis  are
described  here.   Table 5.1 provides  a  checklist  summarizing  the
postsampling procedures.

5. 1  Sample Preparation

    Sample preparation should not be  started  until the required 36-
hour conversion time has elapsed for complete  conversion  of N02~ to
N03~.   When  using  Method  7D  for  relative  accuracy  testing  of
continuous  emission monitors, the sample can be prepared immediately
if the nitrite in the sample  is  quantitated  using  the  procedures
described  in  Subsection  5.2.   The  liquid  level  in  the  sample
container should be checked to  determine  if  sample  has  been lost
during shipment.  If a loss has occured, the appropriate steps should
be  taken to correct for the loss ( see Subsection 5.1.1).  The sample
is  prepared  for  ion  chromatography  by  precipitating  the excess
permanganate  as  manganese  dioxide (Mn02).   A  5%  (v/v)  hydrogen
peroxide (H202)  solution is used to reduce the permanganate to MnO,,.
The Mn02 precipitate is removed by vacuum filtration and the filtered
solution is volumetrically diluted prior to chromatographic analysis.

5.1.1 Sample Loss Determination and Correction - Before preparing the
sample, it must be allowed the full 36 -hour conversion time.  Compare
the  liquid  level  in  the  sample  container  to the  mark  on  the
container.  If  a  noticeable amount of sample has been lost, use the
following procedure for correcting the sample volume:

  1. Mark the new or final level of liquid on the sample container.

  2. Transfer the sample  to  a 1-liter volumetric flask (V    ), and
rinse the container with water.                            soin

  3.  Fill  the  sample  container  with  water to the initial sample
level  marked on the  container  after  sampling.  Transfer the water
to   a  graduated  cylinder,   and  determine  the   original  sample
volume (
  4. Fill  the sample container with water to the final sample level.
Transfer the water to a graduated  cylinder,  and determine the final
sample volume (vsoln )•
                    *f
                    JS
         ^\^f J- Alp               ^\JA,lli
by using Equation 5-1:
5. If V  .    is less than Vgoln ,  correct the sample volume (V    )

-------
                                                   Section No. 3.15.5
                                                   Date July 1, 1986
                                                   Page 2


              Vsoln '  - Vsoln   Vsolni                Equation 5-1
                                v
where:                           soln,.
       i                              i
  V      = sample volume to be used for calculations, ml,
   SO xll
  V  ,   = volumetric flask volume, ml,
   soln
  V      = initial sample volume placed in sample container, ml, and
   so J. n.

  V  n   = final sample volume removed from container, ml.
   soln.p


    6.  Both the corrected and uncorrected values should be submitted
in the test report to the Agency.

5.1.2 Permanganate Precipitation and Filtration  - After the required
36-hour conversion period for the  sample  has elapsed and the sample
container  has  been  checked  for  sample  loss,  the  sample can be
quantitatively transfered to a 1-liter  volumetric  flask.   (If  the
correction  for  sample loss has already been made, the sample should
already  be  in  a  1-liter  volumetric flask.)   The  procedure  for
precipitating the excess permanganate is as follows:

    1.  Dilute the sample in  a  1-liter  volumetric flask (V     ) to
volume with reagent water, and mix well.

    2.  Take a 50-ml aliquot  (V )  of  the  sample  from the 1-liter
volumetric flask,  and  transfer3 the  aliquot to a 250-ml Erlenmeyer
flask  containing  a  Teflon-coated  stirring  bar.    If   the   NO
concentration is low, a 100-ml aliquot may be taken to  increase  th§
instrument response.

    3.  Stir the sample as fast as possible without splashing  any of
the sample out of the flask.

    4.  Add a 5% H202  solution in 5-ml portions while stirring until
the permanganate color disappears.

    5.  Stop stirring and allow the precipitated manganese dioxide to
settle.  If  the  solution is clear, then enough H2<32 has been added.
If the permanganate color persists in the solution, then continue the
H202  addition  in  5-ml  portions until a clear solution is produced
after settling.

    6.  Assemble the Buchner funnel and  filter  flask.  The spout of
the Buchner funnel may be fitted with a  length  of  Teflon tubing to
minimize  the  probability   of  sample  loss  by  aspiration  during
filtration.

-------
                                                   Section No. 3.15.5
                                                   Date July 1, 1986
                                                   Page 3

    7.  Place a piece of GF/C filter  paper (or an equivalent type of
filter paper) in the Buchner  funnel.   Wet the paper with water,  and
seal the filter by applying a vacuum to the flask.

    8.  Quantitatively transfer the precipitated  sample  solution to
the filter, and filter the solution.  Wash the Erlenmeyer  flask  and
the solid material on the  filter  with water four times, and collect
the washings with the filtered solution.

    9.  Quantitatively transfer the filtered solution from the filter
flask to a 250-ml volumetric  flask  (vvJ*   Dilute  to  volume  with
water.

    10.  Prepare a reagent  blank by repeating steps 2 through 9 on a
diluted  sample  of  the  alkaline-permanganate  absorbing  solution.
Dilute 60 ml of the absorbing solution to 100 ml with  water, and use
50 ml in step 2.

5.2 Sample Analysis by Ion Chromatography

    For Method 7D, the basic  components and the operation of the ion
chromatograph  are the same as for Method 7A.  Discussions of the ion
chromatograph  can  be  found  in  both Section  3.15.2  and  Section
3.14.2.  The analyst  should  be  familiar with the operator's manual
for his particular ion chromatograph  system.   In  this section, the
preparation  of  calibration  standards, the use of quality assurance
audit samples,  the  analysis  procedure,  and the data reduction and
reporting are described.

5.2.1  Preparation of Calibration Standards - The accuracy of the ion
chromatographic analysis, as in any analysis, depends directly on the
accuracy of the prepared  calibration  standards.   The use of proper
pipetting  procedures,  described in Method 7A, Section 3.14.2, and a
properly dried,  reagent  grade  standard  are  necessary  to  obtain
quality results from  the  analysis.   The  preparation  of  the N0^~
calibration standards is as follows:                               3

    1.  Dry approximately 15 g of potassium nitrate (KNO~) in an oven
at 105  to 110 C for 2 hours.   (Sodium  nitrate  can  also  be  used
provided  the  difference  in the formula weight is considered in the
subsequent  calculations.)   Allow  the  dried  KNO-  to cool to room
temperature in a desiccator before weighing.

    2.  Calibrate an analytical balance  with  a 5-g Class-S calibra-
tion weight  and  a  10-g  Class-S calibration weight to within 2 mg.
Accurately weigh 9 to 10 g of dried KN03 to within 0.1 mg.

    3.  Dissolve  in  reagent  water,  and  dilute  to_ 1  liter in a
Class-A  volumetric  flask.   Calculate  the exact N03~ concentration
using the following formula:

-------
                                                   Section No. 3.15.5
                                                   Date July 1, 1986
                                                   Page 4


        yg NO ~/ml = g of KNCU x 103 x 62'01             Equation 5-2
             d               3         101.1

The stock standard solution should  be stable for 2 months if precau-
tions, such as refrigeration, are  used  to  prevent decomposition by
nitrate-utilizing microorganisms.

    4.  Prepare  a  fresh  working  standard solution for each set of
analyses by pipetting  5  ml of stock standard solution into a 200-ml
Class-A volumetric flask.  Dilute to volume with water.

    5.  Prepare a series of four calibration standards from the fresh
working standard solution.  Pipet 1.0 ml, 3.0 ml, 5.0 ml, and 10.0 ml
into a series of four 100-ml  Class-A  volumetric  flasks.  Dilute to
volume  with  reagent  water.   The  concentration of the_calibration
standards  made from a 9.7823 g KN03/liter (6000  yg  N03~/ml)  stock
standard solution would be 1.5, 4.5, 7.5, and 15.0 yg Nof~/ml.
                                                        o

    The calibration standard concentrations  cited  above are used in
the example used in Figure 5.1, the analytical data form for analysis
of calibration standards.

    The calibration standards for nitrite  quantitation  are prepared
when Method  7D  is  used for relative accuracy testing of continuous
emission monitors.   A  stock  N02~ standard solution is (1) prepared
with NaNO2 of known purity or (2) analyzed before  use.   Do not oven
dry the NaN02.   Dissolve  52.5  mg  of  NaN02 in water and dilute to
volume  in  a 250-ml Class-A volumetric  flasR.   A  series  of  four
calibration_standards with N02   concentrations of 1.4, 4.2, 7.0, and
14.0 yg N02~/ml are prepared By pipetting 10, 3.0, 5.0, and 10 ml of
stock  N02  standard into four 100-ml Class-A volumetric flasks.  The
N02~ calibration standards are diluted to volume with water.

5.2.2   Quality  Assurance  Audit Samples - The quality of analytical
results  can  be assessed by  analyzing  nitrate  standard  solutions
prepared  by an independent laboratory.  For such standard solutions,
or quality assurance  audit  samples, the concentrations are known to
the control agency (the auditor) but are unknown to the analyst.

    Subsection 3.2.6 of the Federal Register  promulgation  of Method
7D  (see  Section 3.15.10) requires the analysis of quality assurance
audit samples  as described in Method 7.  This means that when Method
7D is used to demonstrate compliance with  an  EPA pollutant emission
standard (specified in 40 CFR Part 60),  a  performance audit must be
conducted  on the analytical phase of the method.  Nitrate samples in
glass vials must be obtained  for  this  performance  audit  from the
Quality  Assurance  Management Office at each EPA Regional Office  or
from  the  responsible  enforcement agency.  The addresses of the EPA
Regional Quality Assurance Coordinators  are  shown  in  Table 5.1 of
Section 3.0.5 of this Handbook.

-------
Plant

Date
                                        Location
                                        Analys t
                                                                Section No. 3.15.5
                                                                Date July 1, 1986
                                                                Page 5
                                                               A/o.  3>
                                                                      ate



Standard
identifier
Std 1
Std 2
Std 3
Std 1*


Standard
concentration (x)
(yg/ml N03 )
/.r
4
-------
                                                   Section No. 3.15.5
                                                   Date July 1, 1986
                                                   Page 6

    The  concentration of each audit sample measured by  the  analyst
must  agree  within  10 percent (relative error) of the actual  audit
concentration.   The relative error is calculated using the following
equation:
where
                   RE
                      = Cd -
                                                         Equation 5-3
'a x  100
         C, = Determined audit sample concentration, mg/dscm, and

         C  = Actual audit sample concentration, mg/dscm.
          Cl

5.2.3. Ion Chromatographic Analysis of Calibration Standards, Reagent
Blank,  Field Samples, and Quality Assurance Samples - The  selection
of  the ion Chromatographic conditions depends on the particular  ion
chromatograph  system  available to the analyst.   The  selection  of
eluents for ion  chromatography  depends  on  the method of detection
used.   For  suppressed  ion  chromatography the following conditions
have been used successfully:
   1. A 0.003M NaHCO«/0.0024M  ^
dissolving  1.008  g  of  NaHCO,
diluting to 4 liters.          *
                                       eluent solution is prepared by
                                       1.018
                of  Na2C03  in water  and
   2. The  full-scale  detection  range  is  set at 3
conductance), and a 0.5-ml sample loop is used.
                                                        MHO (units of
   3. A flow rate of  2.5  ml/min  gives  a  NO
approximately 15 minutes depending on the type
                                                   retention  time of
                                                f column used.
   Non-suppressed  ion  chromatography and ion-pairing_chromatography
may also be used  provided  baseline separation of N0~  and SO.
aration  and  detection  of  NO
                                   are  obtained  (see  Figure
                                    sep-
                                   5.2).
Packed-bed  suppression columns  are not recommended for  quantifying
N0«~ when using Method 7D for relative accuracy testing.
   The recommended procedure
as follows:
                              for the ion Chromatographic analysis is
   1. Establish a stable baseline.  Inject  a  sample  of  water, and
observe the chromatogram to see whether  any NO3  elutes.  Repeat the
water injection until N0«  is not observed on tne  chromatogram.  If,
after 5 injections,  a  N03   peak  is  still  seen, the water source
should be checked for contamination.

   2. Inject samples as follows:

      •   calibration standards;

-------
                                            Section No. 3.15.5
                                            Date July 1, 1986
                                            Page 7
                                               Field  Sample:  AP-1
                                            Flow Rate:  1.5 ml/min
                                       Detector:  30 yS  full scale
                                                 Injection: 50 y1
                                            Chart Speed:  1 cm/min
                                               N03  3.3 minutes
                    Inject
Figure 5.2.
Example of a properly documented chromatogram having
adequate baseline separation.

-------
                                                   Section No. 3.15.5
                                                   Date July 1, 1986
                                                   Page 8

      •   field samples, reagent blank, and quality assurance samples
          (duplicate injections); and
      •   finally the calibration standards again.

The injection volumes for all the standards and samples should be the
same.

      3. The  chromatograms  should  be documented  with  the  sample
identification,  injection point, injection volume, nitrate retention
time,  eluent  flow rate, detector sensitivity setting, and  recorder
chart speed (see Figure 5.2).

      4. Manually measure the N03~  peak height or determine the N03~
peak area with an electronic integrator.

5.2.4 Data Reduction and Reporting - The  details  of the data reduc-
tion  procedure are discussed in Section 3.15.6.  The procedures  for
calculating a response  factor  from  the  calibration  standards  by
linear regression  and  for  calculating the % deviation of the known
concentration  of each standard  from  the  predicted  value  are  as
follows:

      1.  Use the analytical data form (see Figure 5.1) for calculat-
ing   the   linear  regression  equation  based  on  the  calibration
standards.

      2.  Record  the   calculated   concentrations   for   the  four
calibration  standards  (x) on the data sheet._ Determine the average
value for the instrument  response (y) for N03~ (peak height or  area
under  the  peak) from the three determinations for each of the  four
calibration standards.

      3.  Plot the average values for the instrument response for the
calibration  standards against the corresponding  calculated  concen-
trations of the calibration  standards.   Draw a smooth curve through
the  points without forcing the curve through zero.  The curve should
be linear.

      4.  Determine the slope (m) and the intercept term (b or I) for
the linear calibration curve by linear regression.   Many  scientific
calculators are capable of performing linear regression.

      5.  It  is  recommended  that  calculations  be   performed  to
determine the percent deviation of the known concentration  value for
each  calibration standard from the concentration  predicted  by  the
calibration  curve.   To  do  this,  first  calculate  the  predicted
concentration  for  each calibration standard (P) using the following
equation:
                                                         Equation 5-4

          isin ~ ^   Average Instrument Response (y) - Intercept (I)
          wu3 ; =         Calibration Curve Slope (m)

-------
                                                   Section No. 3.15.5
                                                   Date July 1, 1986
                                                   Page 9

Then calculate the percent deviation of each calibration standard (x)
from the predicted value as shown in Equation 5-5.

                                                        Equation 5-5

    % Deviation = P (»* N03"/ml> " * <* N03"ml> x 100

                           x (yg N03~/ml)

If  any standard deviates from the standard curve by more  than  +7%,
the problem should be investigated.

    The concentration of the field samples,  the  reagent  blank, and
the quality assurance samples are calculated  by  the  same procedure
used to calculate  the  predicted  values  for  the calibration stan-
dards.   Use the data form shown in Figure 5.3 for  the  analysis  of
field samples.  The procedure is as follows:

    1.  Determine the instrument response  factor  for the sample and
calculate the sample concentration using Equation 5-4.  Calculate the
average value for the two determinations made on each sample.

    2.  It is recommended that calculations be performed to determine
the percent deviation of the concentration measured for each  individ-
ual  sample  from  the average concentration value calculated  for the
two  determinations  made  on  each  sample.   The deviation   can  be
calculated using the following equation:

                                                         Equation 5-6

% Deviation = SamPle Concentration - Average Concentration    10_
                         Average Concentration

    The percent deviation  for  a  sample  should be within 5% of the
average value before the analysis is considered valid.

    The data reduction_procedures described above for  NO ~   analysis
can be  used  for  NO2   analysis  when  using Method 7D tor relative
accuracy testing of continuous emission monitors.

    The main parameters of the analytical  procedures  may be  checked
during  or  after  the analysis, using the posttest  operations  form
(Figure 5.4).

-------
                                                              Section No. 3-15-5
                                                              Date July 1, 1986
                                                              Page 10
Date samples received
Plant
                                     Date samples analyzed
                                             Run number(s)
                                                                 / 2, .
Location
                        MT"
Calibration curve slope (m) 15.
                                      Analyst   S .
                                        Intercept term (I)  — £>, 2,4-3 ^
Field
sample
number
AP-I
AP-I
AP-3
Field
Blank
Analysis
number
1st
2nd
1st
2nd
1st
2nd
1st
2nd
Instrument
response (y)
(mm or area counts)
7? *,*
/fttntr*
fa^ (W
(p~Z-mw
72~to**
/O wn\
Stn^~>
JT w w
Concentration of
analysis samgle
(yg/ml N03 )
S.8
^7
4.?
4-0>
5:3
^.z
0,31
O.W
Average
Concentration of
analysis samgle
(yg/ml N03 )
s = ^. ?r
s = 4^r
s = 5:er
B = 6,3
-------
                                               Section No.  3.15.5
                                               Date July 1,  1986
                                               Page 11
Reagents
Potassium nitrate dried at 105° to 110°C for a minimum
  of 2 hours before use?
Stock standard solution (potassium nitrate) less than 2 months
  old? 	

Sample Preparation
Has liquid level noticeably changed?* 	
  Original volume 	  Corrected volume 	

Analysis
Standard calibration curve prepared?* 	^	
All calibration points within 7 percent of linear calibration
  curve (optional)?	Y€&	
Reagent blanks made from absorbing solution? 	Yes.	
Same injection volume for both standards and samples?    ^
Duplicate sample values agree within 5 percent of their mean?
Audit sample analytical results within 10 percent of true value?
	^££	
All analytical data recorded on checklist and laboratory form?
* Most significant items/parameters to be checked.
                Figure 5.4.  Posttest operations.

-------
                                                               Section No. 3-15.5
                                                               Date July 1, 1986
                                                               Page 12
                     Table 5.1.  ACTIVITY MATRIX FOR SAMPLE ANALYSIS
Characteristics
Acceptance Limits
Frequency and method
   of measurement
Action if
requirements
are not met
Sample Preparation

1.   Conversion

      time



2.   Sample loss
3.  Permanganate
      precipitation
    Permanganate
      filtration
36 hour minimum
Determine sample age
Noticeable amount
Absence of purple
permanganate color
Absence of solids
in the filtrate
Compare sample level
to mark on container
Between each 5 ml
portion of 5% ^2Q2
solution

After filtration is
complete
Hold sample in

container for
36 hours min-
imum time

Correct by pro-
cedure in
Section 5.1.1

Continue adding
5 ml portions
of 5% H202

Refilter
Calibration Stan-
 dards Preparation

1.  ACS grade KNO_
15 g dry KNO-
2.  Stock standard
      solution
3. Calibration
     standards
9 to 10 g of KNO-
accurately weighed
to 0.1 mg; dilute
to 1 liter; store
refrigerated
Standard range to
cover sample range;
maximum allowed
deviation of indi-
vidual standard
from the predicted
value is +1%
(optional!
Oven dry at 105  to
110 C for 2 hours;
cool in desiccator
Calibrate analytical
balance
Use recommended volumes
of stock standard solu-
tion; calculate devia-
tion (optional) using
Equation 5-5
High bias will
occur if stan-
dard contains
moisture; redry
KNO_

Biases will
occur with
poor pipetting
or improper
storage; re-
make standard

Invalid anal-
ysis; remake
and rerun
calibration
standards
(continued)

-------
                                                                Section No. 3.15-5
                                                                Date July 1. 1986
                                                                Page 13
Table 5.1.  (continued)
Characteristics
Acceptance Limits
Frequency and method
   of measurement
   Action if
   requirements
   are not met
Ion Chromatograph
 Analysis

1. Sample injection
    device
2. Sample analysis
3. Chromatogram
     documentation
   Quality
    assurance
Absence of KNO,
on chromatogram of
water injection

Individual sample
replicates within
5% of average
(optional)

Include sample
identification,
injection point,
injection volume,
NO,  retention
time, eluent flow
rate, detector
sensitivity
setting, and
speed chart

Analytical results
must be within 10%
of actual value
Inject reagent water
up to four times
Calculate deviation
(optional) using
Equation 5~6
Visually check
Check water source
for contamination
Invalidate anal-
ysis; reanalyze
samples
Supply missing
"information
Report results to
agency with sample
identification
 Invalidate  anal-
 ysis;  repeat prep-
 aration of  sample;
 prepare new stan-
 dards

-------
                                                   Section No. 3.15.6
                                                   Date July 1, 1986
                                                   Page 1
6.0  CALCULATIONS
    Calculation errors due to procedural or mathematical mistakes can
be a large component of total  system error.  Therefore, it is recom-
mended  that  each  set  of  calculations be repeated or spotchecked,
preferably by a team  member  other  than  the  one who performed the
original calculations.  If a difference greater than  typical  round-
off error is  detected,  the  calculations  should  be  corrected.  A
computer program is advantageous  in reducing calculation errors.  If
a standardized computer program is  used,  the  original  data  entry
should be checked, and  if  differences  are observed, a new computer
run should be made.   Table 6.1 at the end of this section summarizes
the quality assurance activities for calculations.

    Calculations should be carried  at least one extra decimal figure
beyond that of the acquired data,  and  should be rounded after final
calculation to two significant digits  for  each  run or sample.  All
rounding of numbers  should  be performed in accordance with the ASTM
380-76 procedures.  All calculations are then recorded on a form such
as the ones shown in Figure 6.1A and 6.IB, following the nomenclature
list.

6.1  Nomenc1ature

    The following nomenclature is used in the calculations:

    V      =  dry gas volume as measured by the dry gas meter,
              dcm (dcf),

    Y      =  dry gas meter calibration factor, dimensionless,

    P.      =  barometric pressure, mm  (in.) Hg,

    P td   =  standard absolute pressure, 760 mm  (29.92 in.) Hg,

    T      =  average dry gas meter absolute temperature, °K (°R),

    T      =  standard absolute temperature, 293°K (528°R),
     S cQ

    V , f.d\ =  dry gas volume measured  by the dry  gas meter  corrected
      ^    '   to standard conditions,  dscm  (dscf),

    S      =  analysis of sample,  yg N03 /ml,

    B      =  analysis of blank, yg N03 /ml,

    m      =  mass of NO  as N0~ in sample, yg,
                        X      ^

    C      =  concentration of NO  as  N09,  dry basis, mg/dscm
              (Ib/dscf), and

    X      =  CO, correction factor.

-------
                                                   Section No. 3.15.6
                                                   Date July 1, 1986
                                                   Page 2
6.2  Calculations
    The following are the equations  used  with  example  calculation
forms  (Figures  6.1A  and  6.IB) to calculate the  concentration  of
nitrogen oxides in the samples.

6.2.1  Sample Volume - Calculate the sample volume on a dry  basis at
standard conditions  (760  mm (29.92 in.) Hg and 293 K (528  R) using
Equation 6-1.

                    T     p            v  p              Equation 6-1
    V / 4.,-x -  V XY  std   bar =  K.XY  m  bar
     m(std)     m		     1	

where:               m     std             m

     X  =  correction factor for C00 collection,    100	
                                                 100 - %C09(v/v),
     KI =  0.3858    K  for metric units, or
                   mmHg

     K. =  17.64   °R   for English units.
                  in. Hg

6.2.2  Total yg N0~ Per Sample - Calculate  the  total  \i g of NO2 per
sample using Equation 6-2.
        m  = (S - B) 250 x      x 46'01 = 3710 (S - B)

                            50    62.01

                                                         Equation 6-2
where:

      250  =  volume of prepared sample, ml,

    46.01  =  molecular weight of NO ~,

    62.01  =  molecular weight of NO ~,

     1000  =  total volume of KMnO. solution, ml, and

       50  =  aliquot KMnO^ / NaOH solution, ml.


6.2.3  Sample Concentration - Calculate the sample concentration on a
dry basis at standard conditions using Equation 6-3.

              C = K, 	5	
                      Vm                                 Equation 6-3
where:
             -3
    K2  =  10   mg/vig for metric units, or


    K   =  2.205 x 10~9 IbVyg f°r English units.

-------
                                                Section No. 3.15.6
                                                Date July 1, 1986
                                                Page 3

                           Sample Volume




 bar = _^^_'^_ /__ in.  Hg,  T  =  j?" "5 £> . $~~°R



v i +A\ ' 17-64 X  Y Vm  Pbar = _/_._/_ 0  £" dscf         Equation 6-1
  / 4.^x
 m(std)
                         Tm
                     Total y g NO2 Per Sample


S = S_._/ _^_ yg/mi,  B  =  0_*_3 7_ yg/ml


m =  3710 (S-B)=  /  f   ^^   6ygof NO0           Equation  6-2
                      "" ~~  """" " ~""™"  ^—         ^



                          Sample Concentration


C = 2.205 x 10"9	   = _£/_£ &_ b_ x 10~5 Ib/dscf

                   m(std)                               Equation  6-3



                       Sample Concentration in ppm


ppm N02 = 8.375 x  10  C  =	3_ 5_ 2.^ ppm N02          Equation  6-4
  Figure 6.1A.   Nitrogen oxide calculation form (English units).

-------
                                                Section No.  3.15.6
                                                Date July 1,  1986
                                                Page 4
                           Sample Volume
v  =0.0  2
 m       —
J_
                         m3, y  =
                                             ,  X
 bar
          ±_ ?•  nun Hg, T  =  ^  f  ^.
          ~^^^~ ^^-M»          III   '^^^™  — L1- "l^ •" ^"^~ ^^^^
V / 4.^,x = 0.3858 X Y Vm Pbar = 0.0
 m(sta)              - ^ -
                         m
                                       /  3  /  dscm   Equation 6-1
                      Total ug N02 Per  Sample
S =
                      B = J?-_2 f_ vg/ml
m = 3710 (S - B) =
                                        N0
                                    Equation  6-2
C = 10
      4-3     m
           'm(std)
                       Sample Concentration
=  G  3  t> •_/_ mg N02/dscm
                   Sample Concentration  in  ppm
                                                       Equation 6-3
ppm N02 = 0.5228 C = _  5_ ^_  2- ppm
                                     NO,
                                    Equation 6-4
  Figure 6.IB.  Nitrogen oxide  calculation form (metric units).

-------
                                       Section No. 3.15.6
                                       Date July 1,  1986
                                       Page 5
Table 6.1.  ACTIVITY MATRIX FOR CALCULATIONS
Characteristics
Sample volume
calculation
Sample mass
calculation
Sample concen-
tration
Calculation
check
Document and
report re-
sults
Acceptance limits
All data available;
calculations correct
within round-off error
As above
As above
Original and checked
calculations agree
within round-off error
All data available;
calculations correct
within round-off error
Frequency and method
of measurement
For each sample, exam-
ine the data form
As above
As above
For each sample, per-
form independent cal-
culations
For each sample, exam-
ine the data form
Action if
requirements
are not met
Complete the
data, or void
the sample
As above
As above
Check and
correct all
data
Complete the
data, or void
the sample

-------
                                                   Section No.  3.15.7
                                                   Date July 1,  1986
                                                   Page 1
7.0  MAINTENANCE
    The normal use  of emission-testing equipment subjects it to cor-
rosive gases, extremes in temperature, vibration, and shock.  Keeping
the equipment in good operating order over an extended period of time
requires knowledge of the equipment and  a  program  of routine-main-
tenance  which  is  performed quarterly or after 2830 L (100 ft )  of
operation, whichever is greater.  In addition to the quarterly  main-
tenance,  a  yearly  cleaning of the entire meter box is recommended.
Maintenance procedures for the various  components  are summarized in
Table 7.1  at  the  end of the section.  The following procedures are
not required, but are recommended to increase the reliability  of the
equipment.

7.1  Pump

    In the present commercial sampling train, several types of  pumps
are used; the  two'  most  common are the fiber vane pump with in-line
oiler  and  the  diaphragm  pump.   The  fiber  vane  pump requires a
periodic check of the oiler Jar.  Its contents should be translucent;
the oil should be changed if not translucent.  Use the oil  specified
by the manufacturer.   If  none is specified, use SAE-10 nondetergent
oil.   Whenever the fiber vane pump  starts  to  run  erratically  or
during the yearly disassembly, the head  should  be  removed  and the
fiber  vanes  changed.   Erratic  operation of the diaphragm pump  is
normally  due  to  either  a  bad diaphragm (causing leakage)  or  to
malfunctions of the  valves,  which  should  be  cleaned  annually by
complete disassembly.

7.2  Dry Gas Meter

    The dry gas meter should be checked for excess oil  or  corrosion
of the components by  removing  the  top  plate  every 3 months.  The
meter should be  disassembled  and all components cleaned and checked
whenever  the  rotation  of the dials is erratic, whenever the  meter
will  not  calibrate  properly over the required flow rate range, and
during the yearly maintenance.

7.3  Rotameter

    The rotameter should be disassembled and cleaned according to the
manufacturer's  instructions using only recommended  cleaning  fluids
every 3 months or upon erratic operation.

7.4  Sampling Train

    All  remaining  sampling  train  components  should  be  visually
checked every  3  months  and  completely disassembled and cleaned or
replaced yearly.  Many items, such as quick  disconnects,   should  be
replaced   whenever  damaged  rather   than   checked   periodically.
Normally, the best  procedure  for maintenance in the field is to use
another entire unit such as  a  meter  box,  sample box, or umbilical

-------
                                                   Section No. 3.15.7
                                                   Date July 1, 1986
                                                   Page 2

cord (the  hose  that  connects  the sample box and meter box) rather
than replacing individual components.

7.5  Ion chromatograph

    Maintenance activities and schedules  for  ion chromatographs are
make and model  specific.   It  is  therefore  recommended  that  the
analyst  consult  the  operator's manual for instructions relative to
maintenance practices and procedures.

    Guard columns, while not required,  are  recommended for use with
the ion chromatograph in order to extend column lifetime.

-------
                                                                Section No.  3.15-7
                                                                Date July 1. 1986
                                                                Page 3
            Table 7.1.  ACTIVITY MATRIX FOR EQUIPMENT MAINTENANCE CHECKS
  Apparatus
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Sample train
 control con-
 sole
No erratic behavior
Routine maintenance
performed quarterly;
disassemble and clean
yearly
Replace parts
as needed
Fiber vane pump
In-line oiler free of
leaks
Periodically check
oiler jar; remove
head and change fiber
vanes
Replace as
needed
Diaphragm pump
Leak-free valves
functioning properly
Clean valves during
yearly disassembly
Replace when
leaking or mal-
functioning
Dry gas meter
No excess oil, corro-
sion, or erratic rota-
tion of the dial
Check every 3 mo. for
excess oil or corrosion
by removing the top
plate; check valves and
diaphragm yearly and
whenever meter dial runs
erratically or whenever
meter will not calibrate
Replace parts
as needed, or
replace meter
Rotameter
Clean and no erratic
behavior
Clean every 3 mo. or
whenever ball does not
move freely
Replace
Sampling train
No damage
Visually check every
3 mo.; completely dis-
assemble and clean or
replace yearly
If failure
noted, use an-
other entire
meter box,
sample box,
or umbilical
cord
Ion chroma-
 tograph
See owner's manual
See owner's manual
See owner's
manual

-------
                                                   Section No. 3.15.8
                                                   Date July 1, 1986
                                                   Page 1
8.0  AUDITING PROCEDURE
   An  audit  is an independent assessment of data quality.  Indepen-
dence is achieved if the individual(s) performing the audit and their
standards and equipment are different from the regular field team and
their standards and equipment.  Routine quality assurance checks by a
field team are necessary  to generate good quality data, but they are
not part of the auditing  procedure.   Table  8.1  at the end of this
section summarizes the quality assurance functions for auditing.

   Based on the results of collaborative tests  '  '    of  Method  7,
two specific performance audits are recommended:

   1.  Audit of the analytical phase of Method 7D.

   2.  Audit of data processing.

It is suggested that a systems audit be conducted as specified by the
quality  assurance  coordinator  in  addition  to  these  performance
audits.   The  two  performance  audits  and  the  systems  audit are
described in detail in Subsections 8.1 and 8.2, respectively.

8.1  Performance Audits

   Performance audits are made to evaluate quantitatively the quality
of data produced by the  total measurement system (sample collection,
sample analysis, and data processing).  It is recommended that audits
be  performed by the responsible control  agency  once  during  every
enforcement source test.  A source  test  for enforcement comprises a
series  of  runs  at  one  source.   The  performance  audit  of  the
analytical phase is subdivided into two steps:   (1)  a pretest audit
which is optional, and (2) an  audit during the field sampling and/or
analysis  phase  which  is required.  No audit is recommended at this
time for the sample collection phase.

8.1.1  Pretest  Audit  of  Analytical  Phase (Optional) - The pretest
audit described  in  this  section  can  be  used  to  determine  the
proficiency of the analyst, the  quality  of  the  standardization of
solutions in the Method 7D analysis, and the ability  to  perform the
computations correctly.  It should  be performed at the discretion of
the agency auditor, the laboratory supervisor,  source  test company,
or quality assurance officer.  The analytical phase of Method 7D  can
be  audited  with  the  use  of  aqueous potassium or sodium  nitrate
samples  provided to the testing laboratory  before  the  enforcement
source test.  Aqueous potassium  or  sodium  nitrate  samples  may be
prepared by the procedure described in Section 3.15.5 for calibration
standard preparation.

   The pretest audit provides the opportunity for the testing labora-
tory to check the accuracy of its  analytical  procedure.  This audit
is especially recommended for a  laboratory with little or no experi-
ence  with  the  Method  7D  analysis  procedure  described  in  this
Handbook.

-------
                                                   Section No. 3.15.8
                                                   Date July 1, 1986
                                                   Page 2

   As  an  alternative  to preparing their own audit  samples  for  a
pretest  audit, a testing laboratory may make a request 30 days prior
to the time of the  planned  pretest  audit  to  EPA's  Environmental
Monitoring Systems  Laboratory,  Quality  Assurance  Division, Source
Branch,  Mail Drop 77A, Research Triangle Park, North Carolina  27711
for  known quality control samples.   These  EPA  audit  samples  are
aqueous  potassium nitrate samples and  not  aqueous  sodium  nitrate
samples.                                                       ,

   The relative error for  each  of  two  samples should be within 10
percent of true  value.   The relative error (RE) is an indication of
the bias that may be associated with the  analytical  phase of Method
7D.  Calculate RE using Equation 8-1.

               RE _ Cd " Ca   x 10Q                      Equation 8-1

                       Ca

where:

    C, = determined audit sample concentration, mg/dscm, and

    C  = actual audit sample concentration, mg/dscm.

8.1.2  Audit  of  Analytical  Phase during Field Test (Required) - As
stated in Sections  3.3.9  and 4.4 of 40 CFR 60, Appendix A, Method 7
(49 FR  26522,  06/27/84),  when  the  method is used for enforcement
testing, the analyst must analyze two audit  samples  along  with the
field samples.   The  testing  laboratory should notify the responsi-
ble agency requiring the  performance  test  of the intent to test at
least 30 days prior  to  the  enforcement source test.  The responsi-
ble agency  will  provide two audit samples to be analyzed along with
the field samples from the enforcement source test.   The  purpose of
this   audit  is  to  assess  the  data  quality at the time  of  the
analysis.  If EPA is the agency requiring the performance  test,  the
testing  laboratory  should  notify the Quality Assurance  Management
Office in the respective EPA Regional Office.  The  addresses  of the
EPA Regional Quality Assurance Coordinators are shown in Table 5.1 of
Section 3.0.5 of this Handbook.

    The two audit samples and the compliance  samples  should be con-
currently analyzed  in  the  same manner to evaluate the technique of
the analyst, standards preparation,  and  computation skills.  (Note:
It is recommended that known  quality  control  samples  be  analyzed
prior to the compliance  and  audit  sample  analysis to indicate any
problems.   One source of these samples is the Source  Branch  listed
previously  in  Subsection  8.1.1.)   The  same  analyst,  analytical
reagents, and analytical system  must  be  used  both  for compliance
samples  and  the  EPA  audit samples;  if  this  condition  is  met,
auditing  of  subsequent compliance analyses for the same enforcement
agency  within  30 days may not be required.  An audit sample set may
not  be used to validate different sets of compliance  samples  under

-------
                                                   Section No. 3.15.8
                                                   Date July 1, 1986
                                                   Page 3

the  jurisdiction of different  enforcement  agencies,  unless  prior
arrangements are made with both enforcement agencies.

    Calculate  the concentrations  in  mg/dscm  using  the  specified
sample  volume in the  audit  instructions.   (Note:   Indication  of
acceptable results may be obtained immediately by reporting the audit
results in mg/dscm and compliance results  in total mg N02 /sample by
telephone  to  the  responsible  enforcement  agency.)   Include  the
results of both audit samples, their  identification numbers, and the
analyst's  name  with  the  results  of  the compliance determination
samples  in  appropriate  reports to the EPA regional office  or  the
appropriate  enforcement  agency.   Include  this   information  with
subsequent compliance analyses for the same enforcement agency during
the 30-day period.

    The  concentration of each audit sample measured by  the  analyst
must agree within 10 percent  of  the  actual concentrations.  If the
10-percent specification is not met, reanalyze the compliance samples
and audit samples, and  include  initial and reanalysis values in the
test report.

    Failure to meet the 10-percent specification  may require retests
until the audit problems are resolved.  However, if the audit results
do not affect the compliance or noncompliance  status of the affected
facility, the Administrator  may  waive  the  reanalysis requirement,
further audits, or retests and accept  the  results of the compliance
test.   While  steps  are being taken to resolve audit analysis prob-
lems, the Administrator  may also choose to use the data to determine
the compliance or noncompliance  status  of  the  affected  facility.
Other  applications  of  Method  7D  (i.e., Performance Specification
Tests) should follow agency recommended or required procedures.

8.1.3   Audit of Data Processing -g Calculation errors  are prevalent
in  performing this method.  '  '    Data processing  errors  can  be
determined  by auditing the recorded data on the field and laboratory
forms.   The  original  and  audit  (check) calculations should agree
within  round-off error; if not, all of the remaining data should  be
checked.  The data processing  may  also  be audited by providing the
testing  laboratory  with specific data sets (exactly as would appear
in  the  field),  and by requesting  that  the  data  calculation  be
completed   and   that   the    results    be    returned    to   the
agency/organization.  This audit is useful in checking  both computer
programs and manual methods of data processing.

8.2  Systems Audit

    A systems audit is an on-site qualitative  inspection  and review
of the total measurement system (sample collection,  sample analysis,
etc.).    Initially,  a  systems  audit  is  recommended   for   each
enforcement source test, defined here  as  a  series of three runs at
one source.  After the test  team  gains  experience with the method,
the frequency  of  auditing  may be reduced--for example, to once for
every four tests.

-------
                                                   Section No. 3.15.8
                                                   Date July 1, 1986
                                                   Page 4

   The auditor should have extensive  background experience in source
sampling,  specifically  with the measurement system  being  audited.
The functions of the auditor are summarized below:

   1.   Inform  the testing team of the results  of  pretest  audits,
specifying any area(s) that need special attention or improvement.

   2.   Observe  procedures  and  techniques of the field team during
sample collection.

   3.   Check/verify  records  of  apparatus  calibration  checks and
quality control used in the laboratory  analysis  of  control samples
from previous source tests, where applicable.

   4.   Record  the  results  of the audit,  and  forward  them  with
comments to the team management so that appropriate corrective action
may be initiated.

   While on site, the auditor observes the source test team's overall
performance, including the following specific operations:

   1.  Setting up and leak testing the sampling train.

   2.  Preparing the absorbing solution  (if  performed  on-site) and
adding it to the sampling train.

   3.  Collecting the sample at a constant rate less than 500 cc/min.

   4.  Determination of C02 content.

   5.  Sample documentation procedures, sample recovery, and prepara-
tion of samples for shipment.

   Figure 8.1 is a suggested checklist for the auditor.

-------
                                                              Section No. 3-15-8
                                                              Date July 1, 1986
                                                              Page 5
Yes
 /
No
Comment
OPERATION
                           PRESAMPLING PREPARATION

                 1.  Knowledge of process conditions

                 2.  Calibration of pertinent equipment, in particular,
                     dry gas meter and rotameter, prior to each field
                     test
                                ON-SITE MEASUREMENTS

                       3-  Leak-testing of sampling train after sample run

                       4.  Preparation of absorbing solution and its addition
                           to impingers

                       5.  Constant sampling at less than 500 cc/min

                       6.  Measurement of CO- content

                       7.  Recording of pertinent process conditions during
                           sample collection

                       8.  Maintaining the probe at a given temperature
                                  POSTSAMPLING

                       9.  Control sample analysis - accuracy and precision

                      10.  Sample aliquotting techniques

                      11.  Ion chromatographic technique

                            a.  Preparation of standard nitrate samples
                                (pipetting)
                            b.  Calibration factor (+7 % for all standards,
                                optional)
                            c.  Duplicate sample values (+5% of mean, optional)
                            d.  Adequate peak separation

                      12.  Audit results (+10%)

                            a.  Use of computer program
                            b.  Independent check of calculations
                                    COMMENTS
            Figure 8.1.  Method 7D checklist to be used by auditors.

-------
                                                              Section No. 3.15.8
                                                              Date July 1, 1986
                                                              Page 6
                 Table 8.1.  ACTIVITY MATRIX FOR AUDITING PROCEDURE
  Audit
  Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Performance
 audit of
 analytical
 phase
 Measured RE of the
 audit sample should
 be less than 10% for
 both audit results
Frequency;   Once during
every enforcement source
test*
Method; Measure refer-
ence samples and compare
their true values
Review opera-
ting technique
and calibration
check
Data
  processing
  errors
! Original and checked
 calculations agree
 within round-off
 error
Frequency:   Once during
every enforcement source
test*
Method;  Independent
calculations starting
with recorded data on
Figures 4.1 and 5-1
Check and cor-
rect all data
for the audit
period repre-
sented by the
sampled data
Systems
  audit—
  observance
  of technique
 Operational tech-
 nique as described
 in this section of
 the Handbook
Frequency;   Once during
every enforcement source
test* until experience
gained, then every
fourth test
Method;  Observation of
techniques assisted by
audit checklist, Fig. 8.1
Explain to
team their devi-
ations from
recommended
techniques and
note on
Fig. 8.1
*As defined here, a source test for enforcement of the NSPS comprises a series of
 runs at one source.  Source tests for purposes other than enforcement (e.g., a
 research project) may be audited at a lower frequency.

-------
                                                   Section No.  3.15.9
                                                   Date July 1, 1986
                                                   Page 1


9.0  RECOMMENDED STANDARDS FOR ESTABLISHING TRACEABILITY

    To achieve data of desired quality,  two essential considerations
are  necessary:   (1)  the measurement process must be in a state  of
statistical  control  at the time of the  measurement,  and  (2)  the
systematic errors, when combined with the random variation (errors or
measurement),  must result in an acceptable uncertainty.  As evidence
in support of good quality data,  it  is necessary to perform quality
control  checks and independent audits of the measurement process; to
document  these  data;  and   to   use  materials,  instruments,  and
measurement  procedures that can be traced to an appropriate standard
of reference.

    Data must be routinely obtained by  repeat  measurements of stan-
dard  reference  samples  (primary, secondary, and/or  working  stan-
dards)  and the establishment of a condition of process control.  The
working calibration  standards  should  be  traceable to standards of
higher accuracy.

    Class-S weights (made to NBS specifications)  are recommended for
the analytical  balance  calibration.  See Section 3.15.5 for details
on balance calibration checks.

    The  dry  gas  meter must be calibrated against a wet test  meter
that has been verified  by  an  independent  method  as  described in
Section 2.1.1.

    Audit samples (as discussed in Section 3.15.8)  must  be  used to
validate test results for compliance determination  purposes  and are
recommended  as  an independent check on the measurement process when
the method is performed for other purposes.

-------
 10.0    REFERENCE  METHOD*,**

 Method 7D—Determination of Nitrogen
 Oxide Emission* From Stationary Source*

 Alkaline-Permanganate/Ion
 Chromatographic Method

   1. Applicability. Principle. Interferences.
 Precision. Bias, and Stability.
   1.1   Applicability. The method is
 •applicable to the determination of NO.
 emissions from fossil-fuel fired steam
 generators, electric utility plants, nitric acid
 plants, or other sources as specified in the •
 regulations. The  lower detectable limit is
 similar to that for Method 7C. No upper limit
 has been established: however, when using
 the recommended sampling conditions, the
 method has been found to collect NO,
 emissions quantitatively up to 1782 mg/NO,/
 m». as NOi (932 pm NO,).
 '  1.2   Principle. An integrated gas sample is
 extracted from the stack and collected in
 Hlkaline-potassium permanganate solution:
 NO, (NO+NOi)  emissions are oxidized to
 NOi-. Then N0>- is analyzed by ion
 chroma tography.
   1.3   Interferences. Possible interferences
 are SOs and NH>. High concentrations of SOi
 could interfere because SO» consumes MnO<-
 (as does NO,) and. therefore, could reduce
 the NO, collection efficiency. However, when
 sampling emissions from a coal-fired electric
 utility plant burning 2.1-percent sulfur coal
 with no control of SOi emissions, collection
 efficiency was not reduced, in fact
 calculations show that sampling 3000 ppm
 SO) will reduce the MnO«- concentration by
 only 5 percent if all the SOi is consumed in
 the first impinger.
   NH> is slowly oxidized to NO>- by the
 absorbing solution. At 100 ppm NH> in the
 gas stream, an interference of 6 ppm NO, (11
mg NOt/m1) was  observed when  the sample
was analyzed 10 days after collection.
Therefore, the method may not be applicable
to  plants using NH» injection to control NO,
emissions unless  means are taken to correct
the results. An equation has been developed
to  allow quantitation of the interference and
is discussed in Citation 4 of the bibliography.
  1.4  Precision and Bias. The method does
not exhibit any bias relative to Method 7. The
within-laboratory relative standard deviation
for a single measurement was approximately
6 percent at 200 to 270 ppm NO,.
  1.5  Stability. Collected samples are stable
for at least 4 weeks.
  2. Apparatus.
  2.1  Sampling and Sample Recovery. The
sampling train is  the same as in Figure 7C-1
of Method 7C. Component parts are the same
as in Method 7C.  Section 2.1.
2.2.7
2-2.8
2.2.9
      Section  No.  3.15.10
      Date  July  1,   1986
      Page  1


  2.2  Sample Preparation and Analysis.
  2.2.1  Magnetic Stirrer. With 25- by 10-mm
Teflon-coated stirring bars.
  2.23.  Filtering Flask. 500-ml capacity with
sidearm.
  2.2.3  Buchner Funnel. 75-mm ID. The
spout equipped with a 13-mm ID by 90-mm
long piece of Teflon tubing to minimize
possibility of aspirating sample solution
during filtration.
  Z2.4  Filter Paper. Whatman GF/C. 7.0-cm
diameter.
  2.2.5  Stirring Rods.
        Volumetric Flask. 250-mI.
        Pipettes. Class A.
        Erlenmeyer Flasks. 250-ml.
        Ion Chroma lograph. Equipped with
an anion separator column to separate NOt-.
a H* suppressor, and* necessary auxiliary
equipment Nonsuppressed and other forms
of ion  chroma tography may also be used
provided that adequate resolution of NOj- is
obtained. The system must also be able to
resolve and detect NOi-.
  3. Reagents.
  Unless otherwise indicated, all reagents
should conform to the specifications
established by the Committee on Analytical
Reagents of the American Chemical Society,
where such specifications are available:
otherwise, use the best available grade.
  3.1  Sampling.
  3.1.1  Water. Deionized distilled to
conform  to ASTM specification D 1193-74.
Type 3 (incorporated by reference— see
§ 60.17).
  3.1.2  Potassium Permanganate. 4.0 Percent
(w/w). Sodium Hydroxide. 2.0 Percent (w/w).
Dissolve 40.0 g of KMnO. and 20.0 g of NaOH
in 940  ml of water.
  3.2  Sample Preparation and Analysis.
  3.2.1   Water. Same as in Section 3.1.1.
  3i2  Hydrogen Peroxide. 5 Percent. Dilute
30 percent H>Ot 1:5 (v/v) with water.
  3.2.3   Blank Solution. Dissolve 2.4 g of
KMnOi and 1.2 g of NaOH in 96 ml of water.
Alternatively, dilute 60 ml of KMnO./NaOH
solution to 100 ml.
  3.2.4  KNOj Standard Solution. Dry
KNO> at  110 ' C for 2 hours, and cool in a
desiccator. Accurately weigh 9 to 10 g of
KNOj to  within 0.1 mg. dissolve in water, and
dilute to  1 liter. Calculate the exact NOj-
concentration from the following relationship
ug NOj-/m)=g of KNOi XIO'X •
                          62.01

                          101.10
  *Method  7C  is  reproduced  in  this  section  in  addition  to  Method  7D
  since  the  latter    refers   extensively    to  Method  7C and  Method  7C
  is  not reproduced  elsewhere  in  this  Handbook.
  **  Federal  Register,   Volume 49,  No.   189,  September  27,   1984.

-------
                                                                                     Section  No.   3.15.10
                                                                                     Date  July  1,   1986
                                                                                     Page  2
 This solution is stable for 2 months without
 preservative under laboratory conditions.
   3.2.5   Eluent, 0.003 M NaHCOj/0.0024 M
 Na,CCs. Dissolve 1.008 g NaHCOs and 1.018 g
 NajCOj in water, and dilute to 4 liters. Other
 eiuents capable of resolving nitrate ion from
 sulfate  and other species present may be
 used.
   3.2.6   Quality Assurance Audit Samples.
 This is  the same as in Method 7, section 3.3.8.
 When requesting audit samples, specify that
 they be in the appropriate concentration
 range for Method 7D.
   4. Procedure.
   4.1  Sampling. This is the same as in
 Method 7C. Section 4.1.
   4.2  Sample Recovery. This is the same as
 in Method 7C. Section 4.2.
   4.3  Sample Preparation for Analysis. Note
 the level of liquid in the sample container,
 and determine whether any sample was lost
 during shipment. If a noticeable amount of
 leakage has occurred, the volume lost can be
 determined from the difference between
 initial and final solution levels, and this value
 can then be used to correct the analytical
 result. Quantitatively transfer the contents to
 a 1-liter volumetric flask, and dilute to
 volume.

   Sample preparation can be started 36 hours
 after collection. This time is necessary to
 ensure that all NOr- is converted to NOj-.
 Take a SO-ml aliquot of the sample and
 blank, and transfer to 250-ml Erlenmeyer
 flasks. Add a magnetic stirring bar. Adjust
 the stirring rate to as fast a rate as possible
 without loss of solution. Add 5 percent
 HjOa in increments of approximately S ml
 using a  5-ml pipette. When the KMnO. color
 appears to have been removed, allow the
 precipitate to settle, and examine the
 supernatant liquid. If the liquid is clear, the
 HjO? addition  is complete. If the
 KMnO.  color persists, add more H»Oi, with
 stirring, until the supernatant liquid is clear.
 Note.—The faster the stirring rate, the less
 volume  of HjO» that will be required to
 remove  the KMnO«.) Quantitatively transfer
 the mixture to a Buchner funnel contaiing
GF/C filter paper, and filter the precipitate.
The spout of the Buchner funnel should be
equipped with  a 13-mm  ID by 90-mm long
piece of Teflon tubing. This modification
minimizes the possibility of aspirating sample
solution during filtration. Filter the mixture
into a 500-ml filtering flask. Wash the solid
material four times with water. When
filtration is complete, wash the Teflon tubing,
quantitatively transfer the filtrate to a 250-ml
volumetric flask,  and dilute to volume. The
 sample and blank are now ready for
 NCs analysis.
  4.4 Sample Analysis. The following
 chromatographic conditions are
 recommended: 0.003 M NaHCOj/0.0024 M
 NaiCOi eluent solution. (3.2.5). full scale
 range 3  pMHO: sample loop, 0.5 ml: flow rate,
 2.5 ml/min. These conditions should give a
 NOi- retention time of approximately 15
 minutes (Figure 7D-1).
       /	I
                  1M.M
    Flfim KM.  t«"

  Establish a stable baseline. Inject a sample
of water, and determine if any NOi- appears
in the chromatogram. If NO>- is present,
repeat the water load/injection procedure
approximately five times: then re-inject a
water sample, and observe the
chromatogram. When no NOi- is present, the
instrument is ready for use. Inject calibration
standards. Then inject samples and a blank.
Repeat the injection of the calibration
standards (to compensate for any drift in
response of the instrument). Measure the
NOj peak height or peak area, and determine
the  sample concentration from the calibration
curve.
  4.5  Audit analysis. This is the same as in
Method 7. Section 4.4
  5. Calibration.
  5.1  Dry Gas Metering System (DGM).
  5.1.1  Initial Calibration. Same  as in
Method 0. Section 5.1.1. For detailed
instructions on carrying out this calibration, it
is suggested  that Section 3.5.2 of Citation 3 in
the bibliography  be consulted.
  5.1 2  Post-Test Calibration Check. Same
as in Method 6. Section 5.1.2.
  5.2  Thermometers for DGM and
Barometer. Same as in Method  6.  Section 5.2
and 5.4, respectively.
  5.3  Calibration  Curve for Ion
Chromatograph.  Dilute a given volume  (1.0 ml
or greater) of the KNCs standard solution to a
convenient volume with  water, and use this
solution to prepare calibration standards.
Prepare at least four  standards to cover the
range of the samples being analyzed. Use
pipettes for all additions. Run standards as
instructed in Section 4.4. Determine peak
height or area, and plot the individual values
versus concentration in »ig NCy-/ml. Do not
force the curve through zero. Draw a smooth
curve through the points. The curve should be
linear. With the  linear curve, use linear
regression to determine the calibration
equation.

-------
    6. Calculations.
    Carry out calculations, retaining at least
  one extra decimal figure beyond that of the
  acquired data. Round off figures after final
  calculation.
    6.1  Sample Volume. Dry Basis. Corrected
  to Standard Conditions. Same as in Method
  7C. Section 6.1.
    62  Total Hg NO, Per Sample.
                                                                                   Section  No.  3.15.10
                                                                                   Date  July  1,   1935
                                                                                   Page  3
                  1000    46.01
m-(S-B)X250x    	 x 	
                  50     62.01
Where:
m«=Mass of NO,, as NOi. in sample, fig.
S«= Analysis of sample, fig NCs-/ml.
B«> Analysis of blank, ^g NCS-/ml.
250— Volume of prepared sample, ml.
46.01 - Molecular weight of NOi-.
62.01 «= Molecular weight of NO>-.
1000<=Total volume of KMnO« solution, ml.
50-Aliquot KMnO./NaOH solution, ml.
  6.3  Sample Concentration.
                      m
             C=Kj ———
                                   3710 (S-BT    fEq. 7D-1)
 Where:
 C«= Concentration of NO, as NO>, dry basis,
     mg/dscm.
 V«i««>«=Dry gas volume measured by the dry
     gas meter, corrected to standard
     conditions, dscm.
   6.4  Conversion Factors.
 1.0 ppm NO=1.247 mg NO/m'at STP.
 1.0 ppm NOj-1.912 mg NOj/m'at STP.
 Ift3-2.832xl0-lm'.
   7. Quality Control.
  Quality control procedures are specified in
Sections 4.1 J (flow rate accuracy) and 4.5
(audit analysis accuracy) of Method 7C.
  &.~Bibliogrophy.
  1. Margeson. ).H.. W.J. Mitchell, J.C. Suggs.
and M.R. Midgett. Integrated Sampling and
Analysis* Methods for Determining NO,
Emissions at Electric Utility Plants. U.S.
Environmental Protection Agency. Research
Triangle Park. N.C. Journal of the Air
Pollution Control Association. 32.1210-1215.
1982.
  2. Memorandum and attachment form J.H.
Margeson. Source Branch. Quality Assurance
Division. Environmental Monitoring Systems
Laboratory, to The Record. EPA. March 30.
1983. NHi Interference in Methods 7C and 70.
  3. Quality Assurance Handbook for Air
Pollution Measurement Systems. Volume
III— Stationary Source Specific Methods. U.S.
Environmental Protection Agency. Research
Triangle Park. N.C. Publication No. EPA-600/
4-77-027b. August 1977.
  4. Margeson. J.H.. et al. An Integrated
Method for determining NO. Emissions at
Nitric Acid Plants. Manuscript submitted to
Analytical Chemistry. April 1984.

-------
                                                                Section  No.   3.15.10
                                                                Date   July   1,   1986
                                                                Page   4
Method 7C—Determination of Nitrogen
Oxide Emissions From Stationary Sources

AUialine-Pernianganote/Colorimetric
Method
  1. Applicability, Principle. Interferences,
Precision, Bias, and Stability.
  1.1  Applicability. The method if
applicable to the determination of NO,
emissions from fossil-fuel Tired steam
generators, electric utility plants, nitric acid
plants, or other sources as specified in the
regulations. The lower detectable limit is 13
mg NO./m',  as NO« (7 ppm NO.) when
sampling at 500 cc/min for 1 hour. No upper
limit has been established: however, when
using the recommended sampling conditions,
the method has been found to collect NO,
emissions quantitatively up to 1,782 mg NO,/
m', as NO, (932 ppm NO.).
  1.2  Principle. Ac integrated gas sample is
extracted from the stack and collected in
alkaline-potassium permanganate solution;
NO, (NQ+NOj) emissions are oxidized to
NOT- and NOi-. The NOr- is reduced to
NOr- with cadmium, and the NOr- is
analyzed colorimetrically.
  1J  Interferences. Possible interferences
are SOj  and NH». High concentrations of SO»
could interfere because SO* consumes MnO<-
(as does NOJ and. therefore, could reduce
the NO, collection efficiency. However, when
sampling emissions from i coal-fired electric
utility plant burning 2.1-percent sulfur coal
with no control of SOj emissions, collection
efficiency was not reduced. In fact.
calculations show that sampling 3000 ppm
SO, will reduce the MnO«- concentration by
only 5 percent if all the S0> is consumed in
the first impinger.
  NH> is slowly oxidized to NOr- by the
absorbing solution. At 100 ppm NH» in the
gas stream, an interference of 6 ppn NO. (11
mg NOj/m5) was observed when the sample
was analyzed 10 days after collection.
Therefore, the method may not be applicable
to plants using NH» injection to control NO,
emissions unless means are taken to correct
the results. An equation has been developed
to allow quantilation of the interference and
is discussed in Citation 5 of the bibliography.
  1.4  Precision and Bias. The method does
not exhibit any bias relative to Method 7. The
within-laboratory relative standard  deviation
for a single measurement is 2.8 and 2-fl
percent at 201 and 268 ppm NO,, respectively.
  1.5  Stability. Collected samples are stable
for at least 4 weeks.
  2. Apparatus.
  2.1  Sampling and Sample Recovery. The
Sampling train is shown in Figure 7C-1. and
component parts are discussed below.
Alternative apparatus and procedures are
allowed provided acceptable accuracy and
precision can  be demonstrated.
                                            RESTRICTED omncE WPINGEMS
'ROBE (END PACKED
 WITH GLASS WOOL
                                                      SURCl TANS
                              Figure 7C-1.  110  sampling tr»ir

-------
   2.1.1  Probe. BorosiUcate glass tubing.
 sufficiently heated to prevent water
 condensation and equipped with an to-slack
 or out-stack filter to remove participate
 matter (• plug of glass woo) is satisfactory for
 this purpose). Stainless steel or Teflon tubing
 may also be used for the probe. (Note.
 Mention of trade names or specific products
 does not constitute endorsement by the UJ5.
 Environmental Protection Agency.)
   2.1.2  Impingers. Three restricted-orifice
 glass impingers. having the specifications
 given in Figure 7C-2. are required for each
 sampling train. The impingers must be
 connected in series with leak-free glass
 connectors. Stopcock grease may be used, if
 necessary, to prevent leakage. (The impinjrcrs
 can-be fabricated by « glass blower until they
 become available commercially.)
                          IM/II
           I4IAO.
  DIMINSIONI: M
                            I
                                       11
      f\9ur» K'l. 'fattrictM crlfiu
  2.1 J  Class Wool Stopcock Creese.
Drying Tube. Valve. Pump. Barometer, and
Vacuum Gauge and Rotameter. Same as in
Method 6, Sections 2.1.3. 2.1.4. 2.1.6. 2.1.7,
2.1.6. 2.1.11. and 2.1.12. respectively.
     Section  No.   3.15.10
     Date  July   1,   1986
     Page  5
  2.1.4  Rate Meter. Rotameter. or
equivalent, accurate to within 2 percent at the
selected flow rate between 400 and 500 cc/
min. For rotameters, a range of 0 to 1 liter/
min is recommended.
  2.1.5  Volume Meter. Dry gas meter
capable of measuring the sample volume,
under the sampling conditions of 400 to 500
cc/min for 60 minutes within an accuracy of 2
percent.
  2.1.6  Filter-To remove NO. from ambient
air. prepared by  adding 20 g of a 5-angstrom
molecular sieve to a cylindrical tube. e.g.. a
polyethylene drying tube.
  2.1.7  Polyethylene Bottles. 1-liter, for
sample recovery.
  2.1.8  Funnel and Stirring Rods. For  sample
recovery.
  2.2  Sample Preparation and Analysis.

  2.2.1  Hot Plate. Stirring type with 50- by
10-mm Teflon-coated stirring bars.
  2.2.2  Beakers. 400-. 600-. and 1000-ml
capacities.
  2.2.3  Filtering Flask. 500-ml capacity with
side arm.
  2.2.4  Buchner Funnel 75-mm ID. with   .
spout equipped with a 13-mm ID by 90-mm
long piece of Teflon tubing to minimize
possibility of aspirating sample solution
during filtration.
  2.2.5  Filter Paper. Whatman CF/C. 7.0-cm
diameter.
  2.2.6  Stirring Rods.
  2.2.7  Volumetric Flasks. 100-, 200- or 250-.
500-. and 1000-ml capacity.
  2.2.8  Watch Classes. To cover 600- and
1.000-ml beakers.
  2.2.9  Graduated Cylinders. 50- and 250-ml
capacities.
  2^.10  Pipettes. Class  A
  2.2.11  pH Meter. To measure pH from 0.5
to 12.0
  2-2.12  Burette. 50-ml with a micrometer
type stopcock. (The stopcock is Catalogue
No. 6225-1-05. Ace Class. Inc.. Post Office
Box 996. Louisville. Kentucky 50201.) Place a
glass wool plug in bottom of burette. Cut off
burette at a height of 43 cm from the top of
plug, and have a  glass blower attach a glass
funnel to top of burette such that the
diameter of the burette remains essentially
unchanged. Other means of attaching the
funnel are acceptable.
  2.2.13  Glass Funnel. 75-mm ID at the top.
  2.2.14  Spectrophotometer. Capable of
measuring absorbance at 540 nm. One-cm
cells are adequate.
  2.2.15  Metal Thermometers. Bimetallic
thermometers, range 0 to 150 *C
  £2.18  Culture Tubes. 20- by 150-mm.
Kimax No. 45046.
  2.2.17  Parafllm "M." Obtained from
American Can Company, Greenwich.
Connecticut 06830.
  2.2.18  CO> Measurement Equipment
Same as in Method 3.
  3. Reagents.
  Unless otherwise indicated, all reagents
should conform to the specifications
established by the Committee on Analytical
Reagents of the American Chemical Society,
where such specifications are available;
otherwise, use the best available grade.

-------
    3.1  Sampling.
    3.1.1  Water. Deionized distilled to
  conform to ASTM specification D 1193-74,
  Type 3 (incorporated by reference—see
  § 60.17).
    3.1.2  Potassium Permanganate, 4.0 percent
  (w/w). Sodium Hydroxide. 2.0 percent (w/w).
  Dissolve 40.0 g of KMnO. and 20.0 g of NaOH
  in 940 ml of water.
    3.2  Sample Preparation and Analysis.
    3.2.1  Water. Same as in Section 3.1.1.
    3.2.2  Sulfuric Acid. Concentrated H,SO..
    3.2.3  Oxalic Acid Solution. Dissolve 48 g
  of oxalic acid |(COOH)>-2HiO) in water, and
  dilute to 500 ml. Do not heat the solution.
    3.2.4  Sodium Hydroxide. 0.5 N. Dissolve
  20 g of NaOH in water, and dilute to 1 liter.
    3.2.5  Sodium Hydroxide. 10 N. Dissolve
  40 g of NaOH in water and dilute to 100 ml.
    3.2.6  Ethyienediamine Tetraacetic Acid
  (EDTA) Solution. 6.5 Percent. Dissolve 6.5 g of
  EDTA (disodium salt) in water, and dilute to
  100 ml. Solution is best accomplished by
  using a magnetic stirrer.
    3.2.7  Column Rinse Solution. Add 20 ml of
  6.5 percent EDTA solution to 960 ml of water.
  and adjust the pH to 11.7 to 12.0 with 0.5 N
  NaOH.
    3.2.8  Hydrochloric Acid (HC1). 2 N. Add
  86 ml of concentrated HC1 to a 500-ml
  volumetric flask containing water, dilute to
  volume, and mix well.  Store in a glass-
  stoppered bottle.
    3.2.9  Sulfanilamide Solution. Add 20 g of
  sulfanilamide (melting point 165 to  167 *C) to
  700 ml of water. Add. with mixing, 50 ml
  concentrated phosphoric acid (85 percent),
  and dilute to 1000 ml. This solution is stable
  for at least 1 month, if refrigerated.
    3.2.10  N-(l-Naphthyl)-Ethylenediamine
  Dihydrochloride (NEDA) Solution. Dissolve
  0.5 g of NEDA in 500 ml of water. An aqueous
  solution should  have one absorption peak at
  320 run over the range of 260 to 400 run.

  NEDA. showing more than one absorption
  peak over this range, is impure and should
  not be used.  This solution is stable for at
  least 1 month if  protected from light and
  refrigerated.
    3.2.11  Cadmium. Obtained from Matheson
  Coleman and BelL 2909 Highland Avenue.
  Norwood. Ohio  4521Z as EM Laboratories
  Catalogue No. 2001. Prepare by rinsing in 2 N
  HC1 for 5 minutes until the color is silver-
  grey. Then rinse the cadmium with  water
  until the rinsings are neutral when tested
  with pH paper. CAUTION: Hi is liberated
  during preparation. Prepare in an exhaust
  hood away from any flame.
    3.2.12  NaNOt Standard Solution. Nominal
  Concentration. 100 n g NOr-/ml. Desiccate
  NaNOi overnight. Accurately weigh 1.4 to 1.6
  g of NaNOt (assay of 97 percent NaNOt or
  greater), dissolve in water, and dilute to 1
  liter. Calculate the exact NOr- concentration
  from the following relationship:

                             purity, %           46.01
ug NOi-/ml=g of NaNOtX   	  XlO'X  ——
                               100             o9.01
    Section  No.  3.15.10
    Date  July  1,   1935
    Page  6

  This solution is stable for at least 6 months
  under laboratory conditions.
    3.2.13  KNOj Standard Solution. Dry KNO»
  at 110 *C for 2 hours, and cool in a desiccator.

 Accurately weigh 9 to 10 g of KNQ, to within '
 0.1 mg. dissolve in water, and dilute to 1 liter.
 Calculate the exact NOi- concentration from
 the following relationship:
   NCWml-g of KNO.X10'
62.01

101.10
This solution is stable for 2 months without
preservative under laboratory conditions.
  3.2.14  Spiking Solution. Pipette 7 ml of the
KNOi standard into a 100-ml volumetric
flask, and dilute to volume.
  3.2.15  Blank Solution. Dissolve 2.4 g of
KMnO« and 1.2 g of NaOH in 96 ml of water.
Alternatively, dilute 60 ml of KMnO./NaOH
solution to 100 mi.
  3.2.16  Quality Assurance Audit Samples.
Same as in Method 7, Section 3.3.9. When
requesting audit samples, specify that they be
in the appropriate concentration range for
Method 7C.
  4. Procedure.
  4.1  Sampling.
  4.1.1  Preparation of Collection Train. Add
200 ml of KMnO./NaOH solution (3.1.2) to
each of three impingers. and assemble the
train as shown in Figure 7C-1. Adjust probe
heater to a temperature sufficient to prevent
water condensation.
  4.1.2  Leak-Check Procedure. A leak-check
prior to the sampling  run should be carried
out: a leak-check after the sampling run is
mandatory. Carry out the leak-check(s)
according to Method 6. Section 4.1.2.
  4.1.3  Check of Rotameter Calibration
Accuracy (Optional). Disconnect the probe
from the first impinger. and connect the filter .
(2.1.6). Start the pump, and adjust the
rotameter to read between 400 and 500 cc/
mm. After the  flow rate has stabilized, start
measuring the volume sampled, as recorded
by the dry gas meter (DOM), and the
sampling time. Collect enough volume to
measure accurately the flow rate, and
calculate the flow rate. This average flow
rate must be less than 500 cc/min for the
sample to be valid: therefore, it is
recommended that the flow rate be checked
as above prior to each test
  4.1.4 Sample Collection. Record the initial
DGM reading and barometric pressure.
Determine the sampling point or points
according to the appropriate regulations, e.g..
Section 60.46(c) of 40 CFR Part 60. Position
the tip of the probe at the sampling point.
connect the probe to the first impinger, and
start the pump. Adjust the sample flow to a
value between .400 and 500 cc/min.
CAUTION: HIGHER FLOW RATES WILL
PRODUCE LOW RESULTS. Once adjusted.
maintain a constant flow rate during the
entire sampling run. Sample for 60 minutes.
For relative accuracy (RA) testing of
continuous emission monitors, the minimum
sampling time is 1 hour, sampling 20 minutes

-------
at each traverse point. [Note.—When the SO,
concentration is greater than 1200 ppm. the
•ampling time may have to be reduced to 30
minutes to eliminate plugging of the impinger
orifice with MnO,. For RA tests with SO,
greater than 1200 ppm, sample for 30 minutes
(10 minutes at each point)]. Record the DCM
temperature, and check  the flow rate at least
every 5 minutes. At the conclusion of each
run. turn off the pomp, remove probe from the
•tack, and record the final reeding*. Divide
the campie volume by the sampling Ume to
determine the average flow rate. Conduct a
leak-check a* in Section 4.1.2. If a leak is
found, void the test nia, or we procedures
acceptable to the Administrator to adjust the
sample volume for the leakage.
  4.1.5  CO, Measurement. During sampling.
measure the CO, content of the slack gas
near the sampling point using Method 3. The.
single-point grab sampling procedure is
adequate, provided the measurements are
made at least three times—near the start
midway, and before the end of a run and the
average CO, concentration is computed. The
Orsat or Fyrite analyzer may be used for this
analysis.
  4-2  Sample Recovery. Disconnect the
impingers. Pour the contents of the hnpirrgers
into a 1-liter polyethylene bottle using a
funnel and a stirring rod (or other means) to
prevent spillage. Complete the quantitative
transfer by rinsing the impingers and
connecting tubes with water until the rinsings
that are clear to light pink, and add the
rinsings to the bottle. Mix tbe sample, and
mark the solution  level. Seal and identify the
sample container.
  4J  Sample Preparation for Analysis.
Prepare a cadmium reduction column as
follows: Fill the burette (&2.12) with water.
Add freshly prepared cadmium slowly with
tapping until no further settling occurs. The
height of the cadmium column should be 39
cm. When not in use. store the column  under
rinse solution (3.2.7). (Note.—The column
should not contain any bands of cadmium
fines. This may occur if regenerated column
is used and will greatly reduce the column
lifetime.)
  Note the level of liquid in the sample
container, and determine whether any  sample
was lost during shipment If a noticeable
amount of leakage has occurred, the volume
lost can be determined from the difference
between initial and final solution levels, and
this value can then be used to correct the
analytical result. Quantitatively transfer the
contents to a 1-liter volumetric flask, and
dilute to volume.
  Take  a 100-ml aliquot of the sample and
blank (unexposed KMnO./NaOH) solutions,
and transfer to 400-ml beakers containing
magnetic stirring bars. Using a pH meter, add
concentrated H,SO. with stirring until a pH
of 0.7 is obtained.  Allow the solutions to
stand for 15 minutes. Cover the beakers with
watch glasses, and bring the temperature of
the solutions to 50 "C. Keep the temperature
below 60 *C Dissolve 4.8 g of oxalic acid in a
minimum volume  of water, approximately SO
ml. at room temperature. Do not heat the
solution. Add this solution slowly, in
increments, until the KMnO. solution
Section  No.   3.15.10
Date  July  1,   1986
Page  7

 becomes colorless. If the color is not
 completely removed, prepare some more of
 the above oxalic acid solution, and add until
 a colorless solution is obtained. Add an
 excess of oxalic acid by dissolving 1.6 g of
 oxalic acid in SO ml of water, and add 6 ml of
 this solution to the colorless solution. If
 suspended matter is present, add
 concentrated HfSO4 until a clear solution is
 obtained.

  Allow the samples to cool to near room
temperature, being sure that the samples are
still clear. Adjust the pH to 11.7 to 12-f) with
10 N NaOH. Quantitatively transfer the
mixture to a Buchner funnel containing CF/C
filter paper, and filter the precipitate. Filter
the mixture into a SOO-ml filtering flask. Wash
the solid material four times with water.
When filtration is complete, wash the Teflon
tubing, quantitatively transfer tbe filtrate to a
SOO-ml volumetric flask, and dilute to volume.
The samples are now ready for cadmium
reduction. Pipette a 50-ml aliquot of the
sample into a 150-ml beaker, and add  a
magnetic stirring bar. Pipette in 1.0  ml of 6.5
percent EDTA solution, and mix.'
  Determine the correct stopcock setting to
establish a flow rate of 7 to H ml/min of
column rinse solution  through the cadmium
reduction column. Use a 50-ml graduated
cylinder to collect and measure the solution
volume. After the last of the rinse solution
has passed  from the funnel into the burette,
but before air entrapment can occur, start
adding the sample, and collect it in a 250-ml
graduated cylinder. Complete the
quantitative transfer of the sample  to  the
column as the sample pusses  through  the
column. After the last of. the sample has
passed from the funnel into the burette, start
adding 60 ml of column rinse solution, and
collect the rinse solution until the solution
just disappears from the funnel.
Quantitatively transfer the sample  to a 200-ml
volumetric flask (250-ml may be required).
and dilute to volume. The samples are now
ready for NOr-4- analysis. (Note.— Both the
sample and blank should go through this
procedure. Additionally, two  spiked samples
should be run with every group of samples
passed through the column. To do this.
prepare two additional 50-ml aliquots of the
sample suspected to have tbe highest NOr-
concentration, and add 1 ml of the spiking
solution to these aliquots. If the spike
recovery or column efficiency (see  62.1) is
below 05 percent, prepare a new column, and
repeat the cadmium reduction).
  4.4  Sample Analysis. Pipette 10 ml of
sample into a culture tube. (Note.—Some test
tubes give a high blank NOr- value but
culture tubes do not.) Pipette  in 10 ml of
sulfanilamide solution and 1.4 ml of NEDA
solution. Cover the culture tube with
parafilm. and mix the solution. Prepare a
blank in the same manner using the sample
from treatment of the unexposed KMnO«/
NaOH solution (3.1.2). Also, prepare a
calibration  standard to check the slope of the
calibration  curve. After a 10-minuie color
development interval, measure the
absorbance at 540 nm against water. Rend pg
NOr-/ml from the calibration curve. If the

-------
 absoruance it greater than that of the highest
 calibration standard, pipette less than 10 ml
 of sample and enough water to make the total
 sample volume 10 ml. and repeat the
 analysis. Determine the No? concentration
 using the calibration curve obtained in
 Section 5.3.
  4.5 Audit Analysis. This is the same as in
 Method 7, Section 4.4.
  5. Calibration.
  5.1 Dry Gas Metering System (DCM).
  5.1.1 Initial Calibration. Same as in Method
 6. Section 5.1.1. For detailed instructions on
 carrying out this calibration, it is suggested
 that Section 3.5.2 of Citation 4 in the
 bibiography be consulted.
  5.1.2  Post-Test Calibration Check. Same
 •s in Method  & Section 5.1.2.
  5.2  Thermometers for DCM and
 Barometer. Same as in Method 6. Sections 5.2
 and 5.4. respectively.
  5.3  Calibration Curve for
 Spectrophotomeier. Dilute 5.0 ml of the
 NaNCs standard solution to 200 ml with
 water. This solution nominally contains 25 Kg
 NCWml. Use this solution to prepare
 calibration standards to cover the range of
 0.25 to 3.00 fig NOr-/ml. Prepare a minimum
 of three standards each for the linear and
 slightly nonlinear (described below) range of
 the curve. Use pipettes for all additions.
  Run standards and a water blank as
 instructed in Section 4.4. Plot the net
 absorbance vs ngNO»-/ml. Draw a smooth
 curve through the points. The curve should be
 linear up to an absorbance of approximately
 1.2 with a slope of approximately 0.53
 absorbance units/ pig NOi-/ml. The curve
 should pass through the origin. The curve is
 slightly nonlinear from an absorbance of 1.2
 to 1.6.
  6. Calculations.
  Carry out calculations, retaining at least
 one extra decimal figure beyond that of the
 acquired data. Round off figures after final
 calculation.
  6.1 Sample volume, dry basis, corrected to
 standard conditions.
      Section No.   3.15.10
      Date  July  1,   1986
      Page  8

  6.2  Total >ig NOs Per Sample.
  6il  Efficiency of Cadmium Reduction
Column. Calculate this value as follows:
E
            -  y)'2°0
                       "
     s  x  i-Q
                 SOT

                (Eq. 7C-2)
                           =  269.6  (x -  y)
                                     s
m*

                        „
                      -
                                         (£0. 7C-1)
Where:
V.ua>— Dry gas volume measured by the dry
    gas meter, corrected to standard
    conditions, dscm.
Vw«=Dry gas volume as measured by the dry
    gas meter, don.
Ye Dry gas meter calibration factor.
XK Correction factor for CO> collection.
                     100
      Where:
      E= Column efficiency, unitless.
      x= Analysis of spiked sample, fig NO—/inl.
      y= Analysis of unspiked sample, jig NO—/
          ml.
      200= Final volume of sample and blank after
          passing through the column, ml
        9= Concentration of spiking solution, ng
            NOj/ml.
        1.0= Volume of spiking solution added, ml.
        46.01 = us NCs-/»imole.
        62.01= >ig NOs-/fimole.
          6.2.2  Total fig NCv

   (S-B)         500    1000    (2X10T (S-B)
   -  X200X — x  - =  -
     E            50     100          E

                  "(Eq. 7C-3)

       Where:
       m = Mass of NO., as NO,, in sample, >ig.
       S~ Analysis of sample. u.g NCWml.
       B=Analysis of blank. fig NO— /ml.
       500»=Total volume of prepared sample, ml.
       50 =» Aliquot of prepared sample processed
           through cadmium column, ml.
       100= Aliquot of KMnO./NaOH solution, ml.
       1000=Total volume of KMnOJNaOH
           solution ml.
        6.3  Sample Concentration.  .
                                                                                                C-K,
     Where:
     C= Concentration of NO, as NO:, dry basis,
         mg/dscm.
            "  IOO-*COi(v/v)

PtarK Barometric pressure, mm fig.
P^c Standard absolute pressure, 760 mm Hg.
TK«» Average dry gas meter absolute
    temperature, *K.
T^B Standard absolute temperature, 293 'K.
K, .03858 'K/mmHg.
       6.4  Conversion Factors.
     1.0 ppm NO=1.247 mg NO/m' at STP.
     1.0 ppm NOs =1.912 mg NO2/m'at STP.
     1 ftj«2.832xl0'1mj.
       7. Quality Control.
       Quality control procedures are specified in
     Sections 4.1.3 (flow rate accuracy): 4.3
     (cadmium column efficiency): 4.4 (calibration
     curve accuracy); and 4.5 (audit analysis
     accuracy).
       8. Bibliography.         ,
       1. Marge'son. |.H.. W.J. Mitchell. ).C. Sungs.
     and M.R. Mtdgett. Integrated Samplins and
     Analysis Methods for Determining NO,
     Emissions at Electric Utility Plants. U.S.
     Environmental Protection Agency. Research
     Triangle Park, N.C. Journal of the Air
     Pollution Control Association. J21210-1215.
     1982.

-------
                                                                            Section  No.  3.15.10
                                                                            Date  July  1,   1986
                                                                            Page  9
  2. Memorandum and attachment from J.H.
Margeson. Source Branch. Quality Assurance
Division. Environmental Monitoring System*
Laboratory, to The Record. EPA. March 30.
1983. NHi Interference in Methods 7C and 7D.
  3. Margeson. |.H.. ).C Suggs, and M.R.
Midgett. Reduction of Nitrate to Nitrite with
Cadmium. Anal. Chem. 52:1955-57.1960.
  4. Quality Assurance Handbook for Air
Pollution Measurement Systems. Volume
III—Stationary Source Specific Methods.
August 1977. U.S. Environmental Protection
Agency. Research Triangle Park. N.C.
Publication No. EPA-4>00/4-77-027b. August
1977.
  5. Margeson. J.H.. et al. An Integrated
Method for Determining NO, Emissions at
Nitric Acid Plants. Manuscript submitted to
Analytical Chemistry. April 1984.

-------
                                                  Section No. 3.15.11
                                                  Date July 1, 1986
                                                  Page 1
11.0  REFERENCES
      1.  Federal  Register, Volume 49, No. 189, September 27,  1984.
          Method 7D - Determination of Nitrogen  Oxide Emissions From
          Stationary   Sources,  Alkaline-Permanganate/Ion  Chromato-
          graphic Method.

      2.  Margeson,  J. H., J. E. Knoll, M. R. Midgett,,G. B. Oldaker
          III,  K.  R.  Loder,  P.  M. Grohse, and W. F.   Gutknecht.
          Integrated Method for Determining NO    Emissions at Nitric
          Acid Plants.  Analytical Chemistry, §6, 2607, 1984.

      3.  Small,  H.  T.,  S.  Stevens,  and W. C. Bauman.  Novel  Ion
          Exchange  Chromatographic   Method   Using   Conductimetric
          Determination.  Analytical Chemistry, 47, 11:801, 1975.

      4.  Johnson,   E.   L.   and   R.   Stevenson.    Basic  Liquid
          Chromatography.  Varian Associates, Inc., 1978.

      5.  Yost,  R.  W.,  L.  S.  Ettre, and R. D. Conlon,  Practical
          Liquid  Chromatography,  An   Introduction.   Perkin-Elmer,
          1980.

      6.  Smith, F. C., Jr., and R.  C.  Chang.   The Practice of  Ion
          Chromatography.   John  Wiley  and  Sons,  Inc.,  New York,
          1983.

      7.  Stevens, T. S. and M. A. Langhorst. Agglomerated Pellicular
          Anion-Exchange  Columns for Ion Chromatography.  Analytical
          Chemistry, 54, 6:950, 1982.

      8.  Stevens,  T.  S., G. L. Jewett, and R. A. Bredeweg.  Packed
          Hollow   Fiber   Suppressors    for   Ion   Chromatography.
          Analytical Chemistry, 54, 7:1206, 1982.

      9.  Mulik,   J.   D.,  and  E.  Sawicki.   Ion  Chromatography.
          Environmental Science and Technology, 13, 7:804, 1979.

      10. Stevens, T. S., J. C. Davis, and H. Small. Hollow Fiber  Ion
          Exchange  Suppressor  for  Ion  Chromatography.  Analytical
          Chemistry, 53_, 9:14,88, 1981.

      11. Stevens,  T.  S.  Packed  Fibers  and  New  Columns  Speed,
          Simplify   Ion  Chromatography.   Industrial  Research   and
          Development, September 1983.

      12. Gjerde,  D.  T.,  J.  S.  Fritz, and G. Schmuckler.   Anion
          Chromatography  with Low-Conductivity Eluents.  Journal  of
          Chromatography.  186, 509, 1979.

-------
                                            Section No. 3.15.11
                                            Date July 1, 1986
                                            Page 2

13. Margeson, J. H., W. J. Mitchell,  J.  C.  Suggs,  and M. R.
    Midgett.   Integrated  Sampling and  Analysis  Methods  for
    Determining  NO   Emissions  at  Electric  Utility  Plants.
    Journal of the Air Pollution Control Association, 32, 1210,
    1982.

14. Eubanks, D. R., and J. R.  Stillian.  Care  of  Ion Chroma-
    tography Columns.  Liquid Chromatography, 2^, 2:74, 1984.

15. Hamil,  Henry  F.,  et. al.  The Collaborative Study of EPA
    Methods 5, 6, and 7 in Fossil Fuel  Fired Steam Generators.
    Final Report, EPA-650/4-74-013, May 1974.

16. Hamil, H. F., and  R.  E.  Thomas.   Collaborative Study of
    Method  for the Determination of Nitrogen  Oxide  Emissions
    from    Stationary   Sources    (Nitric    Acid    Plants).
    EPA-650/4-074-028, May 1974.

17. Hamil, Henry F.  Laboratory and Field  Evaluations  of  EPA
    Methods 2, 6,  and  7.   Final  Report,  EPA  Contract  No.
    68-02-0626,   Southwest  Research Institute,  San  Antonio,
    Texas, October 1973.

-------
                                                  Section No.  3.15.12
                                                  Date July 1,  1986
                                                  Page 1
12.0  DATA FORMS

    Blank data forms  are  provided  on  the  following pages for the
convenience of the  Handbook user.  Each blank form has the customary
descriptive title centered at the top  of  the  page.   However,  the
section-page  documentation in the top right-hand corner of each page
of  other sections has been replaced  with  a  number  in  the  lower
right-hand  corner that will enable the user to identify and refer to
a similar filled-in form  in  a  text  section.   For  example,  Form
M7D-1.1  indicates  that  the form is Figure 1.1 in Section 3.15.1 of
the Method  7D section.  Future revisions of these forms, if any, can
be documented by 1.2A,  1.2B,  etc.  Twelve of the blank forms listed
below  are  included  in  this  section.   Four  are  in  the  Method
Highlights subsection as shown by the MH following the form number.


Form               Title

1.1                Procurement Log

2.2                Wet Test Meter Calibration Log

2.4A and 2.4B      Dry Gas Meter Calibration Data Form (English and
                   Metric Units)

2.5 (MH)           Pretest Sampling Checks

3.1 (MH)           Pretest Preparations

4.1                Field Sampling Data Form for NO
                                                  X

4.2                Sample Label

4.3                Sample Recovery and Integrity Data

4.4 (MH)           On-Site Measurements

5.1                Analytical Data Form for Analysis of Calibration
                   Standards

5.3                Analytical  Laboratory  Data Form for Analysis  of
                   Field Samples

5.4 (MH)           Posttest Operations

6.1A and 6.IB      Nitrogen Oxide Calculation Form (English and
                   Metric Units)

8.1                Method 7D Checklist to be Used by Auditors

-------
PROCUREMENT LOG
Item description

Qty-

Purchase
order
number

Vendor

Date
Ord.

Rec.

Cost

Disposition

Comments

                     Quality  Assurance  Handbook  M7D-1.1

-------
wet test meter serial  number
WET  TEST  METER  CALIBRATION LOG
                   Date
Range of wet test meter  flow rate
Volume of test flask Vs  = 	
  Satisfactory leak check?
  Ambient temperature  of equilibrate liquid in wet test meter and reservoir
Test
number
1
2
3
Manometer
reading, a
mm H2O



Final
volume (Vf) ,
L



Initial
volume (V^) ,
L



Total
volume, (Vm)
L



Flask
volume (V ) ,
L



Percent
error, °
%



 Must be less than 10 mm  (0.4 in.) H2O.
Calculations:
    = Vf *
 % error = 100  (V  - V  )/V0 =
         (+1%).
                               Signature of  calibration person
                                                                Quality  Assurance Handbook M7D-2.2

-------
                       DRY GAS METER CALIBRATION DATA FORM (ENGLISH  UNITS)
Date
             Calibrated by
                                                    Meter box number
                                                                           Wet test meter number
Barometer pressure,  P  =
                               in. Hg   Dry gas  meter  temperature correction factor
Wet test
meter
pressure
drop
'a
in. UjO



Rota-
ineter
setting
'
ft3/min



Wet test
meter gas
volume
'
o
°F



Average
gas temp
'C °F Time of run (9),d min Average ratio (Yi),6 (Y_ ),f i D expressed as negative number. Volume passing through meter. Dry gas volume is minimum for at least five revolutions of the meter. c The average of td and td if using two thermometers; the actual reading if using one thermometer. d i o The time it takes to complete the calibration run. e With Y defined as the average ratio of volumes for the wet test and the dry test meters, Y^^ = Y +0.02 Y for calibration and YA = Y +0.05 Y for the posttest checks; thus, 'w (td •*" 460°F) [Pm + (pn/13-6)] Vd (tw + 460°F) (Pm) Y, + Y2 + Y3 (Eq. 1) and Y = (Eq. 2) With Y defined as the average ratio of volumetric measurement by wet test meter to rotameter. Tolerance Y = 1 +J0.05 for calibration and Y +0.1 for posttest checks. V 460 F) [pm + (Dm/13.6)J ri 0 (tw + 460°F) (Pm) (Rs) (Eq. 3) and (Eq. 4) Quality Assurance Handbook M7D-2.4A
-------
Date
    DRY GAS  METER CALIBRATION DATA FORM (METRIC UNITS)
Calibrated by  	  Meter box number 	  Wet test meter  number
Barometer pressure, P  =
                   in.  Hg   Dry gas meter temperature correction factor
Wet test
meter
pressure
drop
(Dm>'a
mm H2O



Rota-
meter
setting
Us),
cc/min



Wet test
meter .gas
volume
'
o
°C



Average
gas temp

-------
Plant name
Location _
Operator
                           FIELD  SAMPLING DATA  FORM FOR NO
                        City
                        Date
                        Sample no.
Probe length/material
Meter box no.
                        Probe setting 	
                        Meter factor (Y)
Sampling point location(s)
Rotameter setting 	
Initial leak check?
CO- concentration
(1)
(2)
Bar press mm (in.) Hg
Rotameter check? 	
Final leak check? 	
   (3) 	
avg
Sampling
time,
min














Total
Clock
time
24 h















Dry gas
meter
readings
L (ft3)














Total
Sample flow
rate setting,
cc/min (ft^/min)















Sample volume
metered,,(V )
L (ft3) m














V
m
avg
Percent
deviation, a
%














Avg
dev
Dry gas
meter temp,
°C (°F)














Avg
Percent deviation =   m	m avg x 100   (must be less than 10 percent).
                        V  avg
                                     Quality  Assurance Handbook  M7D-4.1

-------
SAMPLE LABEL
Plant
Site
Date
Front rinse Fi
Back rinse I 	 1 B£
Solution
Volume: Initial
Cleanup by


Samp
Run
-ont filter | 	
ick filter | 	
L
F


City
le type
number
Front solution
Back solution
evel marked
inal





L
—







n
n)
£
«

































    Quality Assurance Handbook M7D-4.2

-------
                SAMPLE RECOVERY AND INTEGRITY DATA
Plant
Sampling location
                        Field Data Checks
Sample recovery personnel 	
Person with direct responsibility for recovered samples
Sample
number
1
2
3
Blank
Sample
identification
number




Date
of
recovery




Liquid
level
marked




Stored
in locked
container




Remarks
Signature of field sample trustee
                      Laboratory Data Checks
Lab person with direct responsibility for recovered samples
Date recovered samples received 	
Analyst
Sample
number
1
2
3
Blank
Sample
identification
number




Date
of
analysis




Liquid
level
marked




Stored
in locked
container




Remarks
Signature of lab sample trustee
                               Quality Assurance Handbook M7D-4.3

-------
          ANALYTICAL DATA FORM FOR ANALYSIS  OF CALIBRATION STANDARDS
 Plant

 Date
                       Location

                       Analyst
Standard
identifier
Std 1
Std 2
Std 3
Std 4
Standard
concentration (x)
(yg/ml N03 )




Integrator Response,
Peak Height or Area
(y), (mm)
1




2




Avg




Predicted
standard
concentration
(P)
(yg/ml NO- )




Deviation
(*)




Equation for Linear Calibration Curve,  Average Response as a Function of Standard
Concentration
    y = mx + b =  (
      )  x
 where:
    y =  instrument response (mm or area count) =
    m = calibration curve slope
            mm or area count

              yg NO ~/ml
    x = standard concentration (y g NO- /ml)  =
    b = I =  intercept term (mm or area count)  =

Predicted Standard Concentration (P)
    P (yg N03"/ml)  =
 Average Instrument Response (y)  - Intercept  (I)
                           Calibration Curve Slope (m)
    P (for first standard) =
                                            yg N03"/ml
Deviation

    Deviation (%)
    Deviation
    (of first set
    of standards)
P (yg NO ~/ml)  - x  (yg N(

          x (yg N03~/mL)
                        x 100# =
                                        Quality Assurance  Handbook M7D-5.1

-------
      ANALYTICAL LABORATORY  DATA FORM FOR  ANALYSIS OF  FIELD SAMPLES
  Date samples received
  Plant
  Location
  Calibration curve slope (m)
               Date samples  analyzed
                         Run number(s)
                Analyst
                  Intercept  term  (I)
Field
sample
number



Field
Blank
Analysis
number
1st
2nd
1st
2nd
1st
2nd
1st
2nd
Instrument
response (y)
(mm or area counts)




Concentration of
analysis sample
(yg/ml NO^ )


.s .

Average
Concentration of
analysis sample
(yg/ml NO- )
S =
S =
S =
B =
Deviation
(*)




Concentration of
Analysis Sample
(Vg NO-"/ml)

Concentration
(of first sample)
Deviation
Deviation
(of first
standard set)
Instrument Response (y) - Intercept  (I)
     Calibration Curve Slope (m)
 Sample Concentration - Average Concentration
          Average Concentration
                    x 100% =
                                      Quality  Assurance Handbook M7D-5.3

-------
             NITROGEN OXIDE CALCULATION FORM (ENGLISH UNITS)
                          Sample Volume
Vm =
Pbar
 dcf, Y = 	
in. Hg, T  =
                            m
X = 	.

  °R
Vm(std) = 17'64 X Y  m  bar '
                        Tm
                       dscf
             Equation 6-1
                    Total yg NO2 Per Sample
S =	ug/ml, B =	yg/ml
m =  3710 (S-B) =	vg of N02
                                    Equation 6-2
                          Sample Concentration
C = 2.205 x 10
              -9    m
                      x 10~5 Ib/dscf
                  V
                   m(std)
                                    Equation 6-3
                      Sample Concentration in ppm
ppm N02 = 8.375 x 10  C =  .	ppm N02
                                    Equation 6-4
                                 Quality Assurance Handbook M7D-6.1A

-------
V  • 0.0
 m
            NITROGEN OXIDE CALCULATION FORM (METRIC UNITS)



                           Sample Volume
                         m3, Y =
>bar	mm Hg, Tm = _	._ °K
Vm(std) = °*3858 x Y
                               0.0
dscm     Equation 6-1
                         m
                      Total yg N02 Per Sample
          	 ug/ml,   B =	wg/ml
m = 3710 (S-B) =	vgof N02
                                                        Equation 6-2
C = 10
      -3
             m
           'm(std)
                       Sample Concentration
                    	 mg N02/dscm
         Equation 6-3
                   Sample Concentration in ppm
ppm NO2 = 0.5228 C =	ppm
                                     NO,
         Equation 6-4
                                   Quality Assurance Handbook M7D-6.1B

-------
                    METHOD 7D CHECKLIST TO BE USED BY .AUDITORS
 Yes
No
Comment
OPERATION
                                  PRESAMPLING PREPARATION

                         1.  Knowledge of process conditions

                         2.  Calibration of pertinent equipment, in particular,
                            dry gas meter and rotameter, prior to each field
                            test
                 5.

                 6.

                 7.
                                 ON-SITE MEASUREMENTS

                            Leak- testing of sampling train after sample run

                            Preparation of absorbing solution and its addition
                            to impingers

                            Constant sampling at less than 500 cc/min

                            Measurement of CO- content

                            Recording of pertinent process conditions during
                            sample collection
                        8.  Maintaining the probe at a given temperature
                                   POSTSAMPLING

                        9.  Control sample analysis - accuracy and precision

                        10.  Sample aliquotting techniques

                        11.  Ion  chromatographic technique

                              a.  Preparation of standard nitrate samples
                                  (pipetting)
                              b.  Calibration factor (+7 % for all standards,
                                 optional)
                              c.  Duplicate sample values (+5# of mean, optional)
                              d.  Adequate peak separation

                        12.  Audit results  (+10%)

                              a.  Use  of  computer program
                              b.  Independent check of calculations
                                      COMMENTS
OU.S. GOVERNMENT PRINTING OFFICE: 1991-548-18 Tt 0522
                                           Quality Assurance Handbook M7D-8.1

-------