United States         Environmental Monitoring Systems
             Environmental Protection     Laboratory
             Agency           Research Triangle Park NC 27711


             Research and Development    EPA/600/4-77/027b August 1988
vvEPA      Quality Assurance
             Handbook for
             Air Pollution
             Measurement
             Systems:
             Volume III. Stationary
             Sources  Specific
             Methods

             Section 3.16

-------
   August 1988
                             Volume III

                        Table of Contents


Section                                          Pages    Date


             Purpose and Overview of the Quality            3       1-04-85
             Assurance Handbook

    3 0       General Aspects of Quality Assurance for
             Stationary Source Emissions
             Testing Programs

    3.0.1     Planning the Test Program                   12       5-01-79
    3.0.2     General Factors Involved in Stationary          9       5-01-79
             Source Testing
    3.0.3     Chain-of-Custody Procedure for Source          7       5-01-79
             Sampling
    3.04     Procedure for NBS-Traceable Certification      14       6-09-87
             of Compressed Gas Working Standards
             Used for  Calibration and Audit  of
             Continuous Source Emission Monitoring
             (Revised  Traceability Protocol No. 1
    305     Specific Procedures to Assess                 58       9-23-85
             Accuracy of Reference Methods Used for
             SPNSS
    306     Specific Procedures to Assess                 14       9-23-85
             Accuracy of Reference Methods Used foi
             NESHAP
    307     Calculation and Interpretation of              14      11-05-85
             Accuracy for Continuous Emission
             Monitoring Systems (CEMS)
    308     Audit Materials Available from                 /      1 1 04-85
             U.S.E.P.A.
    309     Continuous Emission Monitoring              47       6-01-86
             Systems (CEMS) Good Operating
             Practices
    3010    Guideline for Developing Quality              11      1126-85
             Control Procedures for Gaseous
             Continuous Emission Monitoring
             Systems

    3 1       Method 2  Determination of Stack Gas
             Velocity and Volumetric Flow Rate

    31  1     Procurement  of Apparatus and Supplies        lb       1 15-80
    312     Calibration of Apparatus                      21       1 15-80
    313     Presamplmg Operations                       7       1-15-80
    31.4     On-Site Measurements                      12       1 15-80
    3.1:5     Postsampling Operations                      3       1-15-80
    3.1.6     Calculations                                 4       1 15-80
    3 1 7     Maintenance                                1       1 15-80
    3.1.8     Auditing Procedure                           5       1 15-80
    3 1 9     Recommended Standards for Establishing       1       1 15-80
             Traceability
    3.1.10    Reference Method                           11       1-15-80
    3111    References                                  2       1 15-80
    3112    Data Forms                                 8       1 15-80

    3 2       Method 3  Determination of Carbon
             Dioxide. Oxygen Excess Air  and Dry
             Molecular Weight

    3.2.1     Procurement of Apparatus 
-------
                                                          August 1988
               Table of  Contents (continued)


Section                                          Pages    Date


3.2.4         On-Site Measurements                      12       1-15-80
    3.2.5     Postsampling Operations                      2       1-15-80
    3.2.6     Calculations                                 3       1-15-80
    3.2.7     Maintenance                                1       1-15-80
    3.2.8     Auditing Procedure                           5       1-15-80
    3.2.9     Recommended Standards for                   i       1-15-80
             Establishing Traceability
    3.2.10    Reference Method                            3       1-15-80
    3.2.11    References                                  .1       1-15-80
   .3.2.12    Data Forms                                  6       1-15-80

    3.3       Method 4—Determination of Moisture
             in Stack Gases

    3.3.1     Procurement of Apparatus and Supplies         9       1-15-80
    3.3.2     Calibration of Apparatus                     19       1-15-80
    3.3.3     Presampling Operations                      7       1-15-80
    3.3.4     On-Site Measurements                 .     10       1-15-80
    3.3.5     Postsampling Operations                      4       1-15-80
    3.3.6     Calculations         .                        8       1-15-80
    3.3.7     Maintenance                                3       1-15-80
    3.3.8     Auditing Procedure                           4       1-15-80
    3.3.9     Recommended Standards for                   1       1-15-80
             Establishing Traceability
    3.3.10    Reference Method                            5       1-15-80
    3.3.11    References                                   1       1-15-80
    3.3.12    Data Forms                                 14       1-15-80

    3.4       Method 5—Determination of Paniculate
             Emissions from Stationary Sources

    3.4.1     Procurement of Apparatus and Supplies        15       1-15-80
    3.4.2     Calibration of Apparatus                     22       1-15-80
    3.4.3     Presampling Operations                     20       1-15-80
    3.4.4     On-Site Measurements                      19       1-15-80
    3.4.5     Postsampling Operations                     15       1-15-80
    3.4.6     Calculations                                10       1-15-80
    3.4.7     Maintenance                                 3       1-15-80
    3.4.8     Auditing Procedure                            7       1-15-80
    3.4.9     Recommended Standards for                   1       1-15-80
             Establishing Traceability
    3.4.10    Reference Method                            6       1-15-80
    3.4.11    References                     •              2       1-15-80
    3.4.12    Data Forms                                 21       1-15-80

    3.5       Method 6—Determination of Sulfur
             Dioxide Emissions from Stationary Sources

    3.5.1     Procurement of Apparatus and Supplies          6       5-01-79
    3.5.2     Calibration of Apparatus                       6       5-01-79
    3.5.3     Presampling Operations                  .     3       5-01-79
    3.5.4     On-Site Measurements                        7       5-01-79
    3.5.5     Postsampling Operations                       7       5-01-79
    3.5.6     Calculations                                  2       5-01-79
    3.5.7     Maintenance                                 1       5-01-79
    3.5.8     Auditing Procedure                            6       9-23-85
    3.5.9     Recommended Standards for                   1       5-01-79
             Establishing Traceability
    3.5.10    Reference Method                             4       5-01-79
    3.5.11    References                                   1       5-01-79

-------
    August 1988
               Table of  Contents (continued)


Section                                          Pages    Date


   3.5.12   Data Forms                                 13       5-01-79

   3.6      Method 7—Determination of Nitrogen
            Oxide Emissions from Stationary Sources

   3.6.1     Procurement of Apparatus and Supplies         5       5-01-79
   3.6.2     Calibration of Apparatus                       5       5-01-79
   3.6.3     Presampling Operations                       5       5-01-79
   3.6.4     On-Site Measurements                       8       5-01-79
   3.6.5     Postsampling Operations                      5       5-01-79
   3.6.6     Calculations                                 4       5-01-79
   3.6.7     Maintenance                                1       5-01-79
   3.6.8     Auditing Procedure                           6       9-23-85
   3.6.9     Recommended Standards for                   1       5-01-79
            Establishing Traceability
   3.6.10   Reference Method                            5       5-01-79
   3.6.11   References                                  1       5-01-79
   3.6.12   Data Forms                                 13       5-01-79

   3.7      Method 8—Determination of Sulfuric Mist
            and Sulfur Dioxide Emissions from
            Stationary Sources

   3.7.1     Procurement of Apparatus and Supplies         7       5-01-79
   3.7.2     Calibration of Apparatus                      10       5-01-79
   3.7.3     Presampling Operations                       4       5-01-79
   3.7.4     On-Site Measurements                      10       5-01-79
   3.7.5     Postsampling Operations                      9       5-01-79
   3.7.6     Calculations         •                        6       5-01-79
   3.7.7     Maintenance                                2       5-01-79
   3.7.8     Auditing Procedure                           3       5-01-79
   3.7.9     Recommended Standards for                   1       5-01-79
            Establishing Traceability
   3.7.10   Reference Method                            5       5-01-79
   3.7.11   References                                  1       5-01-79
   3.7.12   Data Forms                                 17       5-01-79

   3.8      Method 10—Determination of Carbon
            Monoxide Emissions from Stationary
            Sources

   3.8.1     Procurement of Apparatus and Supplies        13       1-04-82
   3.8.2    Calibration of Apparatus                      18       1-04-82
   3.8.3    Presampling Operations                       6       1-04-82
   3.8.4    On-Site Measurements                      12       1-04-82
   3.8.5    Postsampling Operations        '"            5       1-04-82
   3.8.6    Calculations                                 3       1-04-82
   3.8.7    Maintenance                                2       1-04-82
   3.8.8    Auditing Procedure                           7       1-04-82
   3.8.9    Recommended Standards for                   7       1-04-82
            Establishing Traceability
   3.8.10   Reference Method                            3       1-04-82
   3.8.11   References                                  2       1-04-82
   3.8.12   Data Forms                                 11       1-04-82

   3.9      Method 13B—Determination  of Total
            Fluoride Emissions from Stationary
            Sources (Specific-Ion Electrode Method)

-------
                                                          August 1988
               Table  of Contents (continued)

Section                                          Pages     Date


   3.9.1     Procurement of Apparatus and Supplies       20        1-04-82
   3.9.2     Calibration of Apparatus                     25        1-04-82
   3.9.3     Presampling  Operations                      6        1-04-82
   3.9.4     On-Site Measurements                      21        1-04-82
   3.9.5     Postsampling Operations                    19        1-04-82
   3.9.6     Calculations                                 7        1-04-82
   3.9.7     Maintenance                               3        1-04-82
   3.9.8     Auditing Procedure                          8        1-04-82
   3.9.9     Recommended Standards for                  1        1-04-82
            Establishing Traceability
   3.9.10   Reference Method                           2        1-04-82
   3.9.11   References                                  1        1-04-82
   3.9.12   Data Forms                                22        1-04-82

   3.10     Method 1 3A—Determination of Total
            Fluoride Emissions from Stationary
            Sources (SPADNS Zirconium Lake
            Method)

   3.10.1   Procurement of Apparatus and Supplies       13        1-04-82
   3.10.2   Calibration of Apparatus                      5        1-04-82
   3.10.3   Presampling  Operations                      3        1-04-82
   3.10.4   On-Site Measurements                       3        1-04-82
   3.10.5   Postsampling Operations                    18        1-04-82
   3.10.6   Calculations                                 7        1-04-82
   3.10.7   Maintenance                               2        1-04-82
   3.10.8   Auditing Procedure                          1        1-04-82
   3.10.9   Recommended Standards for                  1        1-04-82
            Establishing Traceability
   3.10.10  Reference Method                           5        1-04-82
   3.10.11  References                                  1        1-04-82
   3.10.12  Data Forms                                 6        1-04-82

   3.11     Method 1 7—Determination of Paniculate
            Emissions from Stationary Sources
            (In-Stack Filtration Method)

   3.11.1   Procurement of Apparatus and Supplies        9        1-04-82
   3.11.2   Calibration of Apparatus                      2        1-04-82
   3.11.3   Presampling  Operations                      3        1-04-82
   3.11.4   On-Site Measurements                       6        1-04-82
   3.11.5   Postsampling Operations                     1        1-04-82
   3.11.6   Calculations                                 1        1-04-82
   3.11.7   Maintenance                               2        1-04-82
   3.11.8   Auditing Procedure                          2        1-04-82
   3.11.9   Recommended Standards for                  1        1-04-82
            Establishing  Traceability
   3.11.10  Reference Method                          11        1-04-82
   3.11.11  References                                  1        1-04-82
   3.11.12  Data Forms                                 1        1-04-82

   3.12     Method 9--Visible Determination of
            the Opacity Emissions from
            Stationary Sources

   3.12.1   Certification  and Training of Observers         5        4-20-83
   3.12.2   Procurement of Apparatus and Supplies        2        4-20-83
   3.12.3   Preobservation Operations                    2        4-20-83
   3.12.4   On-Site Field Observations                  18        4-20-83
   3.12.5   Postobservation Operations                   2        4-20-83

-------
    August 1988
               Table of Contents (continued)


Section                                          Pages    Date


   3.12.6   Calculations                                 7       4-20-83
   3.12.7   Auditing Procedure                          2       4-20-83
   3.12.8   Reference Method                           5       4-20-83
   3.12.9   References and Bibliography                  1       4-20-83
   3.12.10  Data Forms                                 9       4-20-83

   3.1 3     Methods 6A and 6B—Determinations
            of Sulfur Dioxide, Moisture, and Carbon
            Dioxide Emissions from Fossil Fuel
            Combustion Sources

   3.13.1   Procurement of Apparatus and Supplies        18       7-01-86
   3.13.2   Calibration of Apparatus                     14       7-01-86
   3.13.3   Presampling Operations                      6       7-01-86
   3.13.4   On-Site Measurements                      25       7-01-86
   3.13.5   Postsampling Operations                    15       7-01-86
   3.13.6   Calculations                                9       7-01-86
   3.13.7   Maintenance                                3       7-01-86
   3.13.8   Auditing Procedure                         11       7-01-86
   3.13.9   Recommended Standards for                  1       7-01-86
            Establishing Traceability
   3.13.10  Reference Method                           5       7-01-86
   3.13.11  References                                 2       7-01-86
   3.13.12  Data Forms                                18       7-01-86

   3.14     Method 7A—Determination of Nitrogen
            Oxide Emissions from Stationary
            Sources (Grab Sampling—Ion
            Chromatographic Method)

   3.14.1   Procurement of Apparatus and Supplies        10       7-01-86
   3.14.2   Calibration of Apparatus                     14       7-01-86
   3.14.3   Presampling Operations                      6       7-01-86
   3.14.4   On-Site Measurements                       7       7-01-86
   3.14.5   Postsampling Operations                    11       7-01-86
   3.14.6   Calculations                                6       7-01-86
   3.14.7   Maintenance                                2       7-01-86
   3.14.8   Auditing Procedure                          6       7-01-86
   3.14.9   Recommended Standards for                   1       7-01-86
            Establishing Traceability
   3.14.10  Reference Method                           3       7-01-86
   3.14.11  References                                 2       7-01-86
   3.14.12  Data Forms                                12       7-01-86

   3.15     Method 7D—Determination of Nitrogen
            Oxide Emissions from Stationary
            Sources (Alkaline-Permanganate—Ion
            Chromatographic Method)

   3.15.1   Procurement of Apparatus and Supplies        18       7-01-86
   3.15.2   Calibration of Apparatus                     20       7-01-86
   3.15.3   Presampling Operations                       6       7-01-86
   3.15.4   On-Site Measurements                      10       7-01-86
   3.15.5   Postsampling Operations                    13       7-01-86
   3.15.6   Calculations                                 5       7-01-86
   3.15.7   Maintenance                                 3       7-01-86
   3.15.8   Auditing Procedure                           6       7-01-86
   3.15.9   Recommended Standards  for                   1       7-01-86
            Establishing Traceability

-------
                                                       August 1988
              Table of Contents (continued)


Section                                        Pages     Date


   3.15.10,  Reference Method                         9        7-01-86
   3.15.11   References                               2        7-01-86
   3.15.12   Data Forms                               11        7-01-86

   3.16     Method 18—Measurement of Gaseous
            Organic Compound Emissions by Gas
            Chromatography

   3.16.1    Procurement of Apparatus and Supplies        16        6-30-88
   3.16.2    Calibration of Apparatus                     15        6-30-88
   3.16.3    Presampling Operations                     44        6-30-88
   3.16.4    On-Site Measurements                     33        6-30-88
   3.16.5    Postsampling Operations                    39        6-30-88
   3.16.6    Calculations                               6        6-30-88
   3.16.7    Maintenance                              3        6-30-88
   3.16.8    Auditing Procedure                         8        6-30-88
   3.1 6.9    Recommended Standards for Establishing
            Traceability                               1        6-30-88
   3.16.10   Reference Methods                         22        6-30-88
   3.16.11 ,  References                               5        6-30-88
   3.16.12   Data Forms                               21        6-30-88

-------
                                                                 Section No.  3-16
                                                                 Date June 30,  1988
                                                                 Page 1
                                    Section 3«l6
                METHOD 18 — MEASUREMENT OF GASEOUS ORGANIC COMPOUND
                          EMISSIONS BY GAS CHROMATOGRAPHY
                                      OUTLINE
     Section

SUMMARY
METHOD HIGHLIGHTS
METHOD DESCRIPTION
     1.   PROCUREMENT OF APPARATUS
          AND SUPPLIES
     2.   CALIBRATION OF APPARATUS
     3.   PRESAMPLING OPERATIONS
     4.   ON-SITE MEASUREMENTS
     5.   POSTSAMPLING OPERATIONS
     6.   CALCULATIONS
     7.   MAINTENANCE
     8.   AUDITING PROCEDURE
     9.   RECOMMENDED STANDARDS FOR
          ESTABLISHING TRACEABILITY
    10.   REFERENCE METHODS
    11.   REFERENCES
    12.   DATA FORMS
Documentation
3.16
3.16
3.16.1
3.16.2
3.16.3
3-16.4
3.16.5
3.16.6
3-16.7
3.16.8
3-16.9
3.16.10
3.16.11
3.16.12
Number
of Pages
3
19
16
15
44
33
39
6
3
8
l
22
5
21

-------
                                                                    Section No.  3.16,
                                                                    Date June  30,  1988
                                                                    Page 2

     To  assist   the Handbook user  in applying Section 3-16 to particular   sampling
and analytical techniques, the  following  table provides  a quick cross reference to
each  of  the  subsections  dealing  with  each  of  the   sampling  and  analytical
approaches.

           CROSS  REFERENCES TO  SUBSECTIONS RELATED TO SAMPLING APPROACHES


Activity
Procurementof equipment
Sampling
Analytical
Reagents
Calibration
Sampling equipment
Presampling operations
Survey measurements
Survey preparations
Sample collection
Sample analysis
Interpretation of data
Preparation for test:
equipment
reagents
Packing equipment
On-site Measurements
Sampling
Postsampling Operations
Prepartion of calibration
s tandards
Audit sample analysis
Sample analysis
Calculations
Emission calculations
Auditing Procedures
Performance audits
System audits

All
Methods

l.lpl"
1.2p8
1-3P9

2.1pl

3-lpl
3-2pl
3-3P10
3-4pl3
3-4pl9

3-5P21
3-&P30
3-7P32

4.3P2


S.lpl
5-2p22
5-3P22

6.0pi

S.lpl
8.2p5

Flask
Sampling

1.1P3
1.2p8
1.3P9

2.1pl

3-lpl
3-2p8.
3-3P11
3-4pl7
3- ''Pig

3-5P28
3-6p30
3-7P32

N/A

• ••
5-lpl
N/A
N/A

6.1pl

N/A
N/A
Rigid
Container
Sampling

I.lp4
1.2p8
1-3P9

2.1pl

3-lpl
3-2 P9
3-3P12
3.4pl8
3-4pl9

3-5P28
3.&P30
3-7P32

4-3P2


5-ipl
5.2p22
5-3P22

6.lpl

8.1P4
8.2p5
Direct
Bag
Sampling

l.lpft
1.2p8
1-3P9

2.1pl

3-lpl
3-2p9
3-3P12
3.AP18
3-*pi9

3-5P28
3-6p30
3-7P32

1-3P9


S.lpl
5.2p22
5-2p22

6.1pl

S.lpA
8.2P5
Direct
Interface
Sampling

I.lp6
1.2p8
1-3P9

2.1pl

3-lpl
N/A
N/A
N/A
3-tpl9

3-5P29
3.6P30
3-7P32

4-3P13


S.lpl
5-2p22
5-2p23

6.1pl

8.1p/(
8.2p5
Dilution
Interface
Sampling

1.1P7
1.2p8
1-3P9

2.2P9

3-lpl
N/A
N/A
N/A
S-'ipig

3-5P29
3-&P30
3-7P32

A.3pi*


S.lpl
5-2p22
5.2P25

6.lpi

8. IpS
8.2p5
Adsorption
Tube
Sampling

1.1P3
1.2p8
1-3P9

2.1pl

3-ipl
3-2plO
3-3P12
3-4pi9
3-1pl9

3-5P28
3-6p30
3-7P32

4.3P17


5.1pl2
5-2p22
5.2p26

6.2p2

8. IPS
8.2p5
 "l.lpl" - Means that the  procurement of the  sampling equipment is dicussed  in Section
 beginning on  page 1 of Section 3-16.1.

-------
                                                                 Section No.  3.16
                                                                 Date June 30,  1988
                                                                 Page 3
                                SUMMARY
     Method  18  is  a generic method for measuring gaseous organic compounds.   The
method is based on separating the major gaseous organic components of a gas mixture
with  a  gas chromatograph (GC) and  measuring  the  separated  components  with  a
suitable  detector.   The  gas samples are analyzed immediately as taken  from  the
stack or within a set period of time after being collected in a Tedlar bag or on an
adsorption tube.
     To  identify and quantify the major components, the retention  times  of  each
separated component are compared  with  those  of , known  compounds under identical
conditions.  Therefore, the analyst must identify approximate concentrations of the
organic emission components beforehand.  With this  information,  the  analyst  can
then prepare or purchase commercially available standard  mixtures to calibrate the
GC under physical conditions  identical to those that will be used for the samples.
The  analyst  must  also  have  some presurvey information concerning interferences
arising from other compounds present and indicating the need for sample dilution to
avoid  detector saturation, gas stream filtration to eliminate particulate  matter,
and prevention of sample loss in moisture condensation in the sampling apparatus.
     This method is  structured  to  analyze  approximately 90 percent of the total
gaseous organics emitted from an industrial source.  It does not include techniques
to identify and measure trace amounts of organic compounds,  such as those found in
building air and fugitive emission sources.
     This method will not determine compounds that 1) are polymeric (high molecular
weight), 2) polymerize  before  analysis,  or  3)  have very low vapor pressures at
stack or instrument conditions.
     The range of this  method is from about 1 part per million (ppm)* to the upper
limit governed by GC detector saturation or column overloading.   The  upper  limit
can  be  extended by diluting the stack gases with an inert gas or by using smaller
gas sampling loops.  The sensitivity limit for a compound is defined as the minimum
detectable concentration of that compound, or the  concentration  that  produces  a
signal-to-noise ratio of three  to  one.   The minimum detectable concentration and
limit  of  quantitation  are determined during the presurvey calibration  for  each
compound.                                                           .  _
     The method descriptions given herein  are  based  on the method '  promulgated
October 8, 1983, and  on  corrections  and  additions  published  on  May  30,  1984
(Section 3-16.10).  Revisions  to the method were promulgated February 19, 198? and
these are also described.    Blank  forms  for  recording  data are provided in the
Method Highlights and in Section 3-3-12 for the convenience of Handbook users.

*Note:  Selective  detectors  may  allow  detection and quantitation of far smaller
        concentrations of certain types of gaseous organic compounds.

-------
                                                                 Section No. 3.16
                                                                 Date June 30, 1<588
                                                                 Page 4
                        METHOD  HIGHLIGHTS
     Section 3-16 describes procedures and specifications  for  determining gaseous
organic  compounds  from  stationary sources.  A gas sample is extracted  from  the
stack  at  a  rate proportional to the stack velocity using one of four techniques:
(1) integrated bag sampling, (2) direct interface  sampling, (3) dilution interface
sampling,  and  (4) adsorption tube sampling.  For the first three techniques,  the
sample or diluted sample is introduced directly into the sample  loop  of  the  gas
chromatograph (GC).  The measured sample is then carried  into the GC column with a
carrier  gas where the organic compounds are separated.  The organic compounds then
are each measured quantitatively by  the  GC detector.  The qualitative analysis is
made  by  comparing  the  retention  times  (from  injection to detection)  of known
standards  to  the retention times of the sample compounds.  Once sample  compounds
are  identified,  quantitative  analysis is made by comparing the detector response
for the sample compound to a known quantity of corresponding standard.   Gas samples
collected  on  adsorption  tubes are desorbed from the  adsorption  media  using  a
solvent.  A measured volume  of  the  desorption solution is injected into a heated
injection port where the mixture vaporizes and is carried into the GC column with a
carrier  gas.    The  sample  is  separated into  the  individual  components,   then
qualitatively and quantitatively analyzed in the same manner as a gas sample.
     Because of the number  of  different  combinations of sampling, sample prepar-
ation, calibration procedures,  GC column materials and operating procedures, and GC
detectors covered under this method,  a  set of tables (appearing at the end of the
Method Highlights section) has been developed to assist the tester in selecting and
the test observer in approving an  acceptable  sampling  and  analytical technique.
The compounds listed in these tables were selected based on their current status as
either presently regulated  or  being  evaluated  for future regulations by EPA and
state and local agencies.  Table  A  lists selected organic compounds for Method 18
and provides the user with: (1) the Chemical Abstracts (CA) name,  any synonyms, the
chemical  formula,  the  Chemical  Abstracts  Service  (CAS)  number;   (2)  method
classification and corresponding  references  for more^information;  and (3) whether
EPA currently has an audit cylinder for this compound.
     For a given compound, the sampling  and  analytical  techniques  described  in
Tables  B,  C,  D  and  E  are classified in Table A (Status  of  Selected  Organic
Compounds for Method 18 Sampling and Analysis  Techniques) into one of five classes
as follows:
     1.   Reference  (R).   This  is  a method promulgated by EPA as the compliance
          test method for one or more EPA emission regulations.
     2.   Tentative  (T).   This  is  a method  where  EPA  method  development  is
          completed and documented, but the method has not been promulgated.
     3-   Development (D).  This is a method currently under development by EPA.
     4.   Other  (0).  This is a method developed and documented by an organization
          other than EPA.
     5.   None  (N).  This is a method that has not been developed or validated but
          should work based on experience with similar situations.
     Table B shows all the sampling techniques  described  in  Method 18.  For each
compound, each of the  allowed sampling techniques is rated either:   (1) recommend-
ed, (2) acceptable, (3) theoretical,  (4)  not  recommended,  or  (5) unknown.  The
rating codes for sampling are based on the extent of method validation.   A particu-
lar  sampling technique is rated based on current EPA methodology.  Where EPA meth-
odology does not exist, methodology provided by organizations other than the EPA is
used for  rating.    As  an example on how to use Table B, the rating for benzene is
"T"  for direct interface, "R-12" for Tedlar bags, and "A-9,13" with carbon  disul-

-------
                                                                 Section No. 3.16
                                                                 Date June 30, 1988
                                                                 Page 5

fide  for  adsorbent tubes.  This means that for sampling, there is  no  documented
experience with the direct interface  method,  but  in  theory it could be valid; a
Tedlar bag is recommended as a sampling technique and Reference 12 provides further
description; and charcoal tubes using carbon disulfide as the desorption liquid are
acceptable and References 9 and 13 provide further description.
     Before a final sampling technique is selected,  the source tester will need to
consider the general strengths and weaknesses of each technique.in addition  to the
guidance  provided  in  Table  B.   The  strengths  and weaknesses for the sampling
techniques described in Method 18 are as follows:

Direct Interface or Dilution Interface
Strengths:
Weaknesses:
1.
2.
             3-

             4.
1.
2.
3.
Tedlar Bag

Strengths:
Weaknesses:
2.
3-

4.
1.
Adsorbent Tubes

Strengths:   1.

             2.
             3-
    Can immediately determine if analysis is successful.
    Samples  collected  are in a form that approximates  the  form  in
    stack  emissions  and  minimizes  the time for degradation through
    polymerization, condensation, etc.
    No loss or alteration  in compounds due to sampling since a sample
    collection media (bag or adsorbent) is not used.
    Method of choice for steady state sources when duct temperature is
    below 100°C and organic concentrations . are  suitable  for  the GC
    detector.
    GC must be located at the sampling site.
    A GC equipped with a flame ionization  detector
                                                                  (FID)  cannot  be
                 operated  at a sampling site  if the presence of the H_ flame will
                 be hazardous.
                 Cannot sample proportionally or obtain a time integrated sample.
                 Because results represent only instantaneous  values, they are not
                 totally indicative of non-steady state processes.
    Samples  collected  are  in  a  form that approximates the form in
    stack emissions.
    Samples may be returned to the laboratory for GC analysis.
    Multiple  analyses,  if  necessary,   may  be  performed  on  each
    collected sample.
    Samples can be collected proportionally.
    Unless protected, Tedlar bags  are  awkward and bulky for shipping
    back  to  the  laboratory.   Caution  must be taken to prevent bag
    leaks.
    Stability of compound(s) of interest in Tedlar bags with time must
    be known.  (Maximum permissible storage  time(s)  must be known or
    determined, and must not be exceeded.)
    Polar compounds  generally  should  not  be  collected  due to bag
    adsorption.  There are some exceptions (i.e., ethylene oxide).
    Samples  may  not  be  collected when  the  concentration  of  any
    component present is within explosive limits.        •      . -  •.-.'•


    Samples collected are compact and easy to return to the laboratory
    for analysis.
    Samples may be returned to the laboratory for GC analysis.
    Sample storage time generally can be extended to a week by keeping

-------
                                                                 Section No.  3.16
                                                                 Date June  30,  19*88
                                                                 Page 6

                 samples  at  0 C.   However,  the migration of the collected  com-
                 pound^) through the  charcoal  to  the  backup  portion   may  be a
                 problem.
Weaknesses:  1.  Quantitative recovery percentage of each organic compound  from the
                 adsorbent material must be known.
             2.  Breakthrough sample gas volume for organic compounds as present in
                 the source matrix must be known for the adsorbent material.
             3-  Any  effect  of  moisture  (in  the  stack  gas) on the  adsorbent
                 material collection  capacity  must  be  known.   Moisture  in the
                 sample  above  2 to 3 percent may severely reduce  the  adsorptive
                 capacity.
             4.  Generally,  samples  can  be  collected  conveniently  only  at  a
                 constant rate.
             5.  Samples must be returned to the lab for analysis.
     Table  C lists the recommended GC detectors commonly used with Method  18.  For
each compound, each GC detector is rated either: R - recommended,  A  - acceptable,
T -  theoretical, N - not recommended, or U - unknown.  A particular GC detector is
rated  based  on current EPA methodology.  Where EPA methodology  does  not  exist,
methodology provided by organizations other than the EPA is used for rating.  As an
example  on how to use Table C, the rating for benzene  is  "R-4,12"  for   a  flame
ionization detector (FID), "N" for an electron capture detector (ECD), "T-38" for a
photoionization detector (PID), and "N"  for  an electrolytic conductivity  detector
(ELCD).  This means an FID  is  recommended as the GC detector and References 4 and
12 provide further description, an ECD and  an  ELCD are not recommended, and there
is no documented experience with a PID for benzene,  but  its  use is theoretically
possible based on the ionization potential found in Reference 38.
     Table D presents information on packed columns suitable for GC analysis  of the
selected compounds.   Items  covered  include  column  type  and conditions,  Kovats
Retention Indices (KRI's), if available, and associated literature references.  The
recommended  column  appears  first, the others are acceptable.  Specifically,  any
column  or  condition  that  meets  the Method 18 criteria for peak  resolution is
considered  acceptable.  A particular GC column is recommended based on current EPA
methodology;  where  EPA  methodology  does  not  exist,  methodology  provided by
organizations other than the EPA is used for rating.  Kovats Retention Indices  were
previously used to identify unknown  compounds by comparison of the measured  KRI(s)
for a compound to catalogued KRI's  for  the various columns.  In performing  Method
18, KRI's can be useful  in  selecting  a GC column which will effectively  separate
two (or more) target compounds and/or interferents in an air sample.   In   Table D,
the first number shown for each compound refers to the literature reference for the
column and column conditions suggested for the  recommended  sampling  method (when
available);  the letter(s) associated with this number cross-references the List of
Referenced  Columns  following  Table  D.   Listed  next  for each compound are the
columns  and  conditions  suggested for sampling methods with an acceptable rating;
additional references are provided  for  columns  used for analysis of the  selected
compounds under laboratory conditions.  Supplementary references  provide KRI's (in
parentheses) for certain compounds.   As an example of how to use Table D, the entry
for'  benzene  is  "12-s,   t;  9-k; 13~u; 4-p;  39-d(658), e(557), i(1039),-   h(1104),
v(963)."  This means that the column described in citations s and t in the  Table  D
List of Referenced Columns was specified in the method described  in  Reference 12;
the  column described in citation k in the List of Referenced Columns was specified
in the  method  described in Reference 9; the column described in citation  u in the
List was specified in the method described in Reference 13;  the column described in
citation p in the List was specified in the method  described  in  Reference 4;  and

-------
                                                                 Section No. 3.16
                                                                 Date June 30, 1988
                                                                 Page 7

the columns described in citations d, e, i, h, and v in the List were specified  in
the method described  in  Reference  39-   The  KRI's  for  each  column  under the
conditions given in the List are shown in parentheses.
     The  user  should be aware that interfering  compounds  may  exist  in  source
samples.  Some method development work, using the required presurvey sample, may be
necessary to optimize separation of the compounds of  interest from the interfering
compounds present in  a  source  sample.  As discussed later in Section 3-16.5. any
column that will provide  an  acceptable  resolution  of the compounds can be used.
Only packed columns  are  described  in  Table  D  since  these  are  more commonly
available  to  source  test analysts than capillary  columns.   However,  capillary
columns are permitted in Method 18 for analysis.
     Table E shows the GC calibration preference  for  each  compound  based on the
technique  used  for  sampling.   Where  appropriate,  the  source  of  calibration
standards  is  also  shown.  For each compound, the calibration technique shown  is
rated  either:   (1)  recommended,  (2)  acceptable,  (3)  theoretical,   (4)   not
recommended, or (5) unknown.  A particular  calibration technique is rated based on
current EPA  methodology.   Where  EPA  methodology  does  not  exist,  methodology
provided by organizations other than the  EPA is used for rating.  As an example on
how to use Table E, the rating for benzene is "R-12 (1806)" for gas cylinders,  "N"
for gas injection into a Tedlar bag, "A-12" for liquid injection into a Tedlar bag,
"R-9,13" f°r preparation  of  the  standard  in  desorption  liquid,  and  "T"  for
preparation of the standard  on  an  adsorption  tube followed by desorption.  This
means that gas cylinders assayed and certified against National Bureau of Standards
(NBS) gaseous  Standard  Reference  Material  (SRM)  1806  using  EPA  Traceability
Protocol No. 1 (Reference 5) are recommended as the calibration standard for direct
interface and Tedlar bag samples with Reference 12 providing further information on
the source of the calibration standard; preparation of calibration standards by gas
injection  into  a  Tedlar  bag  is  not recommended;  preparation  of  calibration
standards  by  liquid  injection  into  a Tedlar bag is acceptable and Reference 12
provides  further  information;  preparation  of  calibration   standards   in  the
desorption liquid is the recommended procedure for  use  with  the  adsorption tube
methods described in References 9 and 13; and  preparation of calibration standards
on adsorption tubes followed by desorption is  theoretically  valid  for  use  with
adsorption tube samples.
     Because the number of organic compounds of interest to EPA and state and local
agencies is increasing, and since  EPA  plans  to  conduct  methods development arid
validation studies for many of the organic compounds identified here as well as for
additional  compounds identified in the future, the Method  Highlights  portion  of
Section 3-16 will be updated every two or three years.  As with all other revisions
of Volume III of the Quality Assurance Handbook, those  individuals whose names are
in  the Record Distribution System will automatically receive  the  updated  Method
Highlights section.
     For compounds not currently  listed in the tables, Figure 0.1 may be used as a
general guide in selecting appropriate sampling techniques.  However, any technique
used must meet the criteria described in detail in the subsequent sections.
     The Method Description (Sections 3.16.1 to 3.16.9)  is  based  on the detailed
specifications in the Reference  Method  (Section  3-16.10)  promulgated  by EPA on
October 18, 1983 and corrections and revisions promulgated February 19, 1987- '

1.  Procurement of Apparatus and Supplies

     Section  3-16.1  gives  specifications,  criteria, and design features  for the

-------
                                                 NONPOLAR
                            LOW-TEMPERATURE
                            LOW MOISTURE (< 1%)
             CONCENTRATION
             >1PPMV
CHARACTERIZE
STATIONARY
SOURCE  	
ORGANIC
COMPOUND
EMISSIONS
                                         TEDLAR BAG
                                         ADSORPTION TUBE
                                                 POLAR
                                         DIRECT INTERFACE
                                         DILUTION INTERFACE
                                         ADSORPTION TUBE
                                         DIRECT INTERFACE
                                         DILUTION INTERFACE
                                                 NONPOLAR
                            MEDIUM TEMPERATURE
                            MEDIUM MOISTURE (> 3%)
                                                 POLAR
                     HIGH CONCENTRATION
HIGH TEMPERATURE
HIGH MOISTURE (> 10%)
                                         HEATED TEDLAR BAG
                                         ADSORPTION TUBE
                                         DIRECT INTERFACE
                                         DILUTION INTERFACE
                                         ADSORPTION TUBE (SILICA)
                                         DIRECT INTERFACE
                                         DILUTION INTERFACE
                                         DILUTION INTERFACE
                                                  LOW CONCENTRATION
                            SEMIVOLATILE
             CONCENTRATION
             <1PPMV
                                           MM5
                            VOLATILE
                                           VOS.T
                                                                      DIRECT INTERFACE
                                                                                                       T3 O C/>
                                                                                                       (0 P CD
                                                                                                       oq n- o
                                                                                                       CD CD
                                                                            CD 21
                                                                              O
             Figure 0.1.  General scheme  for selection of appropriate sampling techniques,
                                                                                                         oo

-------
                                                                 Section No. 3.16
                                                                 Date June 30, 1988
                                                                 Page 9

required equipment and materials.  The sampling apparatus for Method 18  is divided
according to  the  different  sampling  approaches.   This section can be used as a
guide for procurement and initial checks of equipment and supplies.   The  activity
matrix (Table 1.1) at  the end of the section is a summary of the details given  in
the text and can be used as a quick reference.

2.  Presampling Preparations

     Section 3-16.2 describes the required calibration procedures for the Method 18
sampling  equipment.   Section  3-16.3 describes, the presampling operations and the
acquisition of supplies and equipment  needed for the sampling.  Preliminary survey
sampling is discussed, including a description of classes of organic  compounds and
the presurvey sampling techniques that  are  generally  used to obtain a sample for
evaluation  purposes.   The  presurvey  sampling  and  analytical methods are  then
described.  Finally, how to select  the  proper  sampling  and analytical equipment
based on the presurvey data is discussed.   The  preliminary survey and presampling
preparation forms (Figures  3-2 and 3-5 of Section 3-16.3) can be used as equipment
checklists.  Suggestions for packing  the  equipment  and supplies for shipping are
given to help minimize breakage and reduce contamination.
     Activity matrices  for  the  calibration  of  equipment  and  the  presampling
operations (Tables 2.1 and 3-1) summarize the activities detailed in the text.

3.  On-Site Measurements                      :

     Section  3-16.4 describes several sampling techniques.  The use of the presur-
vey sample analyses and the sampling matrix tables (Tables A  through  E)  provides
the user with the required information to select the proper sampling  technique.  A
checklist (Figure 4.8) is an easy reference for . field  personnel  rto  use  in  all
sampling activities.   Sampling  and  analyses  using  the direct interface and the
dilution interface methods  are  both  conducted  on-site;  however, to provide for
greater consistency of presentation, the analytical procedures are presented in the
Posttest Operations Section with those for the other sampling techniques.

4.  Posttest Operations

     Section 3-16.5 describes the analytical procedures and the posttest activities
for checking the equipment.  The initial analytical procedure of sample preparation
is shown based  on  the  sampling  technique  used  and includes the procedures for
preparation  of the calibration standards.  The second procedure discussed  is  the
method of introducing a known volume of sample into the  GC and this is followed by
a discussion of GC operations.  The detailed analytical  procedures  can be removed
for use as an easy reference in the  laboratory.   An  activity  matrix (Table 5.1)
summarizes the postsampling operations.
     Section 3-16-6 describes calculations, nomenclature,  and  significant  digits
for  the  data  reduction.   A  programmed  calculator  is  recommended  to  reduce
calculation errors.
     Section 3-16.7  recommends  routine  and preventive maintenance programs.  The
programs are not required, but their use should reduce equipment downtime.

5-  Auditing Procedures

     Section  3-16.8  describes performance and system audits.   Performance  audits
for both  the  analytical phase and the data processing are described.   A checklist

-------
                                                                 Section No. 3-16
                                                                 Date June 30, 1988
                                                                 Page 10

(Figure 8.2) outlines a system audit.
     Section 3-16.9 lists the primary standards to which  the  working standards or
calibration standards should be traceable.

6.  References

     Section 3-16.10 contains  the promulgated Method; .Section 3.16.11 contains the
references used throughout this text; and Section 3-16.12 lists  all the data forms
in Section 3-16 and contains copies of blank  data  forms for those shown completed
in  the  text.   These  may be removed from  the  Handbook,  copied,  and  used  in
performing the method.   Each  form  has a subtitle [e.g., M18-2.5  (Figure 2.5)] to
assist  the  user  in  locating  the  same  completed form in  the  text.   Several
checklists  are not completed in the text and and therefore not reproduced in  this
section.

-------
                                                                               Section  No.  3.16
                                                                               Date June 30,  1988
                                                                               Page 11
     TABLE A.   STATUS OF  SELECTED  ORGANIC  COMPOUNDS FOR METHOD 18
                                                                 SAMPLING AND ANALYSIS TECHNIQUES
                                                                 = = = = = ~ = = = = = = := = = = = = = = = = = ==:=s = = = = = =
Chemical Abstracts Name
                                    Synonyms
                                                         Formula
                                                                     CAS No.
                                                                               |Method| EPA Audit
                                                                               [Class  | Cylinder(ppm)
                                                Alcohols
Methanol
Ethanol
Isopropyl Alcohol
n-Propyl Alcohol
n-Butyl 'Alcohol

Methyl Alcohol
Ethyl Alcohol
2-Propanol
1-Propanol
1-Butanol
Alkanes
CH.O
c J6o
C?H8°
SH8°
CHo°

(67-56-1)
(64-17-5)
(67-63-0)
(71-23-8)
(71-3&-3)

0-6
0-7
0-7
0-8
0-8

30-80
No
No
No
No

Cyclohexane
Hexane
                                                         c6"12
                                                         C6H14
                                                                    (110-82-7) |  0-9  |    80-200
                                                                    (110-54-3) j  0-9  |20-90.1000-3000
                                                Alkenes
Ethylene
Propy lene
| Ethene
1 Propene
l«$
| (74-85-1) |
| (115-07-1) |
N
N
| 5-20,300-700
I 5-20,300-700
                                                 Dienes
1,3-Butadiene
Hexachlorocyclopentadiene
                              |  Butadiene                |
                              |  Perchlorocyclopentadiene|
                            =============================
                                                Aromatic
C.H,
c:ci
 |  (106-99-0)
 j  (77-47-/O
                                                                                 D-io
                                                                                           5-60
                                                                                            No
Benzene
Mesi tylene
Ethylbenzene
Cumene
Xylene (m- , o- . p- )
Toluene
Styrene
2-Naphthylamine
Benzol
1,3,5-Trimethylbenzene

1-Methylethylbenzene
Dime thy Ibenzene
Methylbenzene
E t he ny Ibenzene
2-Naphthylenamine
C6H6
C9H12
C8H 0
r H
9 12
r 'u
8 10
r u
78
C8H8
c?oY
(71-43-2)
(108-67-8)
(100-41-4)
(98-82-8)
(1330-20-7)
(108-88-3)
(100-42-5)
(91-59-8)
T-12
N
0-13
0-13
0-13
0-9,13
0-13
0-14
5-20,6o-4oo
No
No
No
5-20,300-700
5-20,100-700
No
No
                                                Ke tones
Acetone
Methyl Ethyl Ketone
Methyl Isobutyl Ketone
                              |  2-Propanbne
                              I  2-Butanone
                              |  4-Methyl-2-pentanone
                                                          C H,0

                                                          C4H8°
                                                          C6»12°
   (67-64-1)   | 0-15  |     No
   (78-93-3)   I 0-16  j   30-80
   (108-10-1)  I 0-15  j     5-20

                                                Epoxides
Ethylene Oxide
Propylene Oxide
                              I  Epoxy  Ethane             |
                              j  1,2-Epoxy  Propane        j

                                                Sulfides
                                                          C  H.O
                                                          C  H&0
                                                         ==_=====
 |  (75-21-8)   | 0-17  |   5-20
 I  (75-56-9)   I 0-18  j 5-20,75-200
: = = = = =: = = = = = = = = = = = = = = = = ss- = = s = = = = = = = =: =
bis(2-Chloroethyl)  Sulfide
                              |  Mustard  Gas
                                                                     (505-60-2)  |  N
                                                                                            No

-------
                                                                                  Section  No.  3.16
                                                                                  Date  June 30,  1*988
                                                                                  Page  12
TABLE A.  (Continued)
  Chemical Abstracts Name
                                      Synonyms
                                                      |  Formula
                                                                        CAS  No.
|Method|  EPA  Audit
jdass  j  Cylinder(ppm)'
                                                  Halogenated
Ethylidene Chloride
Ethylene Dibromide
Ethylene Dichloride
Propylene Dichloride
1,1,1-Trichloroethane
Bromodichloronethane
Chi o rod ibromome thane
Chloroform
Carbon Tetrachloride
Dichlorodifluoromethane
Methyl Bromide
Methyl Chloride
Methylene Chloride
Tetrachloroethylene
Bromof orm
Trichloroethylene
Trichlorotrifluoroethane
Vinylidene Chloride
Ethyl Chloride
Chlorobenzene
Vinyl Chloride
l,2-Dibromo-3-chloropropane
1,1-Dichloroethane
1 , 2-Dibromoe thane
1 ,2-Dichloroethane
1,2-Dichloro propane
Methyl chloroform


Trichlorome thane
Tetrachloromethane
Freon 12
Bromomethane
Chloromethane
Dichloromethane
Perchloroethylene
Tribromome thane
Trichloroethene
Freon 113
1,1-Dichloroethene
Chloroethane
Monochlorobenzene
Chloroethylene
DBCP
C2H4C12
C2H4Br2
£ H £.
C^H?cl2
C^H^Cl2
CHBTC1
CHBr C?
CHC1
CC1,3
CC1 F
CH Jr2
CHr,Cl
CH~CI_
2 2
CHBr
CUP T
2 ^
C Cl F
C2H2^12
C H Cl
C6H>C1
r u-V i
0 l^1
2 Jn ri
(5-3*-3)
(106-93-4)
(107-06-2)
(78-87-5)
(71-55-6)
(75-27-4)
(124-48-1)
(67-66-3)
(56-23-5)
(75-71-8)
(74-83-9)
(74-87-3)
(75-09-2)
(127-18-4)
(75-25-2)
(79-01-6)
(76-13-1)
(75-35-4)
(75-00-3)
(108-90-7)
(75-01-4)
(96-12-8)
0-19
0-20
T-21
0-22
T-21
N
N
T-23
T-23
0-24
0-25
0-26
T-27
T-21
0-19
T-21
T-21
0-28
0-29
0-19
R-30
0-37
No
5-20,50-300
5-20,100-600
3-20,300-700
5-20
No
No
5-20,300-700
5-20
No
No
No
1-20
5-20,300-700
No
5-20.100-600
5-20
5-20, 100-600
No
5-20
5-30
No
Method Classification Code
  R = Reference -   EPA promulgated method.

      Tentative -
  T
  D
  0 = Other -
  N = None -
                    EPA method development  complete;  EPA  reference  available.
= Development - EPA method currently  under  development.
                Method development  completed  by  organizations other than EPA; reference available.
                No reference  available;  recommendation  based on experience.
     The codes in the method classification  column  describe  the  current status of a sampling and analysis
method for each selected compound.   For example,  the  method  classification code for benzene is: T-12.
This means the current method for benzene  is a  tenative  EPA  method with development complete and the
reference for the method is citation number  12  in Section  3.16.11.

* The availability of EPA audit  cylinders  is shown  in this column where:

   (   ) = Audit  cylinders for this  particular  compound  are  available from EPA in the concentration ranges
           indicated (Reference  4).

     No  = Audit  cylinders for this  particular  compound  are  not  available from EPA. The source tester must
           obtain audit  gas cylinders  from commercial  gas  vendors certified by independent analysis to be
           within 5 percent of the  concentration  claimed by  the  vendor.

-------
                                                                                  Section  No.  3.16
                                                                                  Date  June 30,  1988
                                                                                  Page  13
     TABLE B.    METHOD  18  SAMPLING  TECHNIQUES  FOR SELECTED ORGANIC COMPOUNDS                          . ,
===BS======== ======================================================================= ==========-=======::==
                             |           |         |          Adsorbent Tubes and Desorption Liquid
                             |  Direct    j  Tedlar  j ------------------------------ •=• ------------------------
  Selected Compounds          |lnterface  j   Bag*   | Charcoal*  |    Other **   |    Desorption Liquid***
                                                 Alcohols
Methanol
Ethanol
Isopropyl Alcohol
n-Propyl Alcohol
n-Butyl Alcohol
T
T '
T
T
T
N
N
N
N
N
N
T-7 .
T-7
T-8
T-8
|A-6; Silica Gel|
1 ~ 1
1 ~ 1
j - j
1 ~ 1
Distille.d Water
IX 2-Butanol in CS
IX 2-Butanol in CS2
Carbon Disulfide
Carbon. Disulfide
                                                 Alkanes
  Cyclohexane
  Hexane
= = = s = = = = = = = = = = ss = = = = = = = = = = = = = = =. = = = = = = = = = =
               U     I    T-9
               U     j    T-9
            ===================
                    Alkenes
                                                                                Carbon Disulfide
                                                                                Carbon Disulfide

  Ethylene
  Propy1ene
I     T
I     T
                                            N
                                                      U

                                          u     I     u
                                          ==============
                                                Dienes
                                                         ===============================================
1,3-Butadiene              |     T
Hexachlorocyclopentadiene  |     T
                                                             j A-ll; Porapak
                                                      Carbon Disulfide
                                                          Hexane
                                                                     ===================================
                                                 Aromatic
Benzene
Mesitylene
Ethylbenzene
Cumene
Xylene (m- , o- , p- )
Toluene
Styrene
2-Napthylamine
T | R-12
T | U
T | U
T | U
T | U
T. | U
T | U
T | U
T-9. 13
U
T-13
T-13
T-13
T-9, 13
T-13
T-lA
-
-
-
-
-
-
-
.
Carbon Disulfide
U
Carbon Disulfide
Carbon Disulfide
Carbon Disulfide.
Carbon Disulfide
Carbon Disulfide
Carbon Disulfide

                                                                                   = s = = = = = = = = = =.= = = = = = = = =
                                                 Ketones
  Acetone
  Methyl  Ethyl  Ketone
  Methyl  Isobutyl  Ketone
========s==========s====================
           I    N     |   T-15
           !    N     |     N
           I    N     I   T-15

                    Epoxides
                                                           |A-l6; Ambersorbj
Carbon Disulfide
Carbon Disulfide
Carbon Disulfide

  Ethylene Oxide
  Propylene Oxide
                       T-17
                       T-18
                                                                           |     99:1 BenzeneiCS.,
                                                                              .  Carbon Disulfide
=========================================
                                         =================
                                                 Sulfides

  bis(2-Chloroethyl)  Sulfide  |     T
====================================
                                                                                            (continued)

-------
                                                                                  Section No.  3.16
                                                                                  Date June  30,  1$88
                                                                                  Page 14
  TABLE B.  (Continued)
  Selected Compounds
       |           |         |           Adsorbent  Tubes  and Desorption Liquid
       j  Direct   j  Tedlar |
       (interface |   Bag*  j  Charcoal* |     Other  **
                                                                                 Desorption  Liquid*
                                                  Halogenated
Ethylidene Chloride
Ethylene Dibromide
Ethylene Dichloride
Propylene Dichloride
1,1,1-Trichloroethane
Bromodichloromethane .
Chlorodibromome thane
Chloroform
Carbon Tetrachloride
Dichlorodifluoromethane
Methyl Bromide
Methyl Chloride
Methylene Chloride
Tetrachloroethylene
Broraoform
Trichloroethylene
Trichlorotrifluoroethane
Vinylidene Chloride
Ethyl Chloride
Chlorobenzene
Vinyl Chloride
1 , 2-Dibromo-3-chloropropane
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
U
N-31
R-21
U
R-21
U
U
R-23
R-23
U
U
U
R-27
R-21
U
R-21
R-21
U
U
U
R-30
U
T-19
T-20
T-19
T-22 ,
T-ig
U
U
T-19
T-19 ,
T-24
T-25
T-26
T-32
T-33
T-19
T-34
T-35
T-28
T-29
T-19
T-36
T-37
_
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-

Carbon Disulfide
99:1 BenzeneiMeOH
Carbon Disulfide
15J! Acetone in Cyclohexane
Carbon Disulfide
U
U
Carbon Disulfide
Carbon Disulfide
Methylene Chloride
Carbon Disulfide
Methanol
Carbon Disulfide
Carbon Disulfide
Carbon Disulfide
Carbon Disulfide
Carbon Disulfide
Carbon Disulfide
Carbon Disulfide
Carbon Disulfide
Carbon Disulfide
Carbon Disulfide
Rating Code
 R - Recommended.
 A = Acceptable.
     Theoretical.
                      Based on actual source tests  experience  (sampling
                      valid and is the method of choice  among  Method  18
                                                  and analysis)  this  method  is
                                                  users .
Based on actual source tests or similar source  test  experience  (sampling  and
analysis), this method is valid.  The tester must  evaluate for specific  test.

Method has no documented experience, but in theory could be  valid.
 N = Not Recommended.  Based on actual source  tests  or  similar  source  test  experience  and/or  theory,  this
                      method is invalid.
     Unknown.
                      Method has no documented experience  and  the  theoretical  aspects  of  sampling by  this
                      method are inconclusive.  The  tester must  demonstrate  that  this  sampling method is
                      valid.                .  .
     The rating codes  for  sampling are  based  on  the  extent  of  method validation.   For  example,  the rating
code for benzene is:  T;  R-12;  A-9,13-   This means  that  direct  interface  is  theoretically  possible for
benzene, but no documented experience has  been  found; Tedlar bags are  the recommended  sampling  method for
benzene by the tenative  EPA method referenced in citation  12 in  Section  3.16.11;  and sampling with
charcoal adsorption tubes  is acceptable following  the two  methods referenced  in citations 9 and 13 in
Section 3.16.11.

  * = If condensibles  exist,  use  the  procedure  described in Section 3-l6.ll.
 ** = Solid sorbents  other  than  charcoal  recommended.

*"* = The  recommended desorption solution  is  given  in  this  column.  Analyst should consult the appro-
      priate reference for  details.

-------
                                                           Section No.  3.16
                                                           Date June 30,  1988
                                                           Page 15
 TABLE C. GC DETECTORS  FOR SELECTED ORGANIC COMPOUNDS BY METHOD 18
3s3333::333 = = 333a = = 3:=33333ss3333a3aa::t=33333:: = 3S3333s = 33333 = 333B33sss
                    |             Gas Chromatograph Detector *
1 	
Selected Compounds | FID

Methanol R-4,6
Ethanol R-7
Isopropyl Alcohol R-7
n-Propyl Alcohol R-8
n-Butyl Alcohol R-8

Cyclohexane R-4,9
Hexane R-4,9

Ethylene A-4
Propylene A-4

1 ,3-Butadiene | R-4.10
Hexachlorocyclopentadiene j R-ll

Benzene R-4,12
Mesitylene T
Ethylbenzene R-i;
Curaene R-i;
Xylene (o-,m-,p-) R-4,i;
Toluene R-4,9,1
Styrene R-i;
2-Napthylamine R-l/l
Acetone R-15
Methyl Ethyl Ketone R-4,l6
Methyl Isobutyl Ketone R-4,15
Ethylene Oxide R-4.17
Propylene Oxide R-4,l(
bis(2-Chloroethyl) Sulfide | U
| ECD PID
Alcohols
N T-38
N T-38
N T-38
N T-38
N T-38
Alkanes
N I T-38
N j T-38
Alkenes
N | T-38
N | T-38
Dienes
.41 N | T-38
T | U
Aromatic
N T-38
N T-38
N T-38
N T-38
N T-38
3 N T-38
N T-38
N U
333333333S3S3333S333333S3333SS33:
Ketones
N | T-38
N j T-38
N j T-38
Epoxides
N T-38
N | T-38
Sulfides
| U U
ELCD

N
N
N
N
N

N
N

N
N

N
T

N
N
N
N
N
N
N
N
=============
N
N
N
N
N
U
(continued)

-------
                                                                             Section  No.  3.16
                                                                             Date June 30,  1988
                                                                             Page 16
TABLE C.   (Continued)
Selected Compounds
Gas Chromatograph Detector *
FID
BCD
PID | ELCD
Halogenated
Ethylidene Chloride
Ethylene Dibromide
Ethylene Dichloride
Propylene Dichloride
1,1, 1-Trichlo roe thane
Bromodichlorom ethane
Ch 1 o rod ibromom ethane
Chloroform
Carbon Tetrachloride
Dichlorodifluororae thane
Methyl Bromide
Methyl Chloride
Methylene Chloride
Tetrachloroethylene
Bromoforra
Trichloroethylene
Trichlorotrifluoroethane
Vinylidene Chloride
Ethyl Chloride
Chlorobenzene
Vinyl Chloride
1 , 2-Dibromo-3-chloropropane
R-19
A-A
R-A.21
A-A
Rr-A.21
U
U
R-A.23.
R-A.23
R-2A
R-25
R-26
R-A.27,32
R-A.21
R-19
R-A.21
R-A.21
R-A.28
R-29
R-A.19
R-A.30
U
T
R-20
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
U
U
T-38
T-38
U
U
U
T-38
T-38
N-38
T-38
T-38
T-38
T-38
T-38
T-38
N-38
T-38
T-38
T-38
T-38
R-37 | u
T
T
T
R-22
T
T
T
A-23
A-23
T
T
T
T
T
T
T
T
T
T
T
T
T
Rating Code
 R = Recommended.
                       Based on actual source tests experience  (sampling  and  analysis)
                       this method is valid an is the method of choice  among  Method  18
                       users.
 A = Acceptable.
 T = Theoretical.
Based on actual source tests or similar  source  test  experience
(sampling and analysis),  this method  is  valid.  The  tester  must
evaluate for specific test.

Method has no documented  experience,  but  in  theory  could be
valid.
 N = Not Recommended.   Based on actual  source  tests  or  similar  source  test experience
                       and/or theory,  this  method  is invalid.
 U = Unknown.
                       Method has no documented  experience  and  the heoretical aspects
                       are not conclusive.   The  tester  must  demonstrate  that this
                       detection method is  valid.
     The rating codes for GC detectors  are  based  on  the  detector specified in  the method
that is referenced.   For example,  the  rating  code  for  benzene  is: R-A.12; N; T-38; N.
This means that the  FID is recommended  for  detection of  benzene by both  references A and
12 cited in Section  3-16-H: the  BCD and  the  ELCD  are  not  recommended for benzene; and
detection of benzene with a PID is  theoretically  possible  based on the ionization
potential found in reference J6.

*  The following abreviations are  used  for  the  gas chromatography detectors:
   FID  = Flame Ionization Detector
   ELCD = Electroconductivity Detector
             (Hall Detector)
                                 ECD   =  Electron Capture Detector
                                 PID   =  Photoionization Detector
                                 (with lamps up to 11.7 electron
                                  volts)

-------
                                                                                  Section  No.  3.16
                                                                                  Date June 30, 1988
                                                                                  Page 17
TABLE D.  PACKED COLUMNS SUITABLE FOR ANALYSIS OF  SELECTED  ORGANIC  COMPOUNDS  BY  GAS CHROMATOGRAPHY


  Selected Compounds     '    |  Column Reference,  Type  and  Conditions,  and  Kovats  Retention  Indices*

                                     '                     Alcohols
  Me thanol
  Ethanol
  Isopropyl Alcohol
  n-Propyl Alcohol
  n-Butyl Alcohol
                             6-a;  4-b,  c;  39-d(3?0).  e(331),  f(426)
                             7-g
                             7-g;  39-d(477),  e<396>.  f(5?6),
                             8-j;  39-e(387).  f(533)
                             8-J;  39-d(649)
S3S33Sse38BBecsaBB3aassBB33B8SBaBsasB3B3ssasssB3B33SasB3ssE3SBBB3=
                                                          Alkanes
  Cyclohexane                |  9-k;  4-1;  39-d(66?),  e(5U),  f(6l9)
  Hexane                     I  <)-k;  4-m;  39-d(600),  e(600).  f(600)
                                                          Alkenes
  Ethylene                   |  4-n  '
  Propylene                  j  4-n
                                                          Dienes
  1 , 3-Butadiene              |  lO-o;  4-p.  q,  4l-nnn
  Hexachlorocyclopentadiene  j  11-r
                                                          Aromatic
                               13-u: t-d(869).  e(573).
                               13-u; 4-1(13-18)
                               13-w; 4-x;  39-d(ra = 876,  p = 877.  o =  900),  i(m =  1297.  P  =  1312,  o  =  1353)
                               9-k; 13-w;  A-p;  39-d(76l).  h(1136),  i(1201),  v(1060)
                               i3-w: s
                               14-y
Benzene                    |  12-s, t; 9-k;  13-u;  4-p;  39-d(6s8),  e(557),  1(1039),  h(1104),  v(9&3)
Mesitylene
Ethylbenzene
Cumene
Xylene (o-,n-,p-)
Toluene
Styrene
2-Napthylamine  '

                                                        Ketones

Acetone                    |  15-z; 4-e(380),  f(636),  h(1009),  h(1091)
Methyl Ethyl Ketone        j  16-aa:  4-bb;  39-d(579).  e(476),  f(6A4),  h(1087),  i(1158),  v(92?)
Methyl Isobutyl Ketone     j  15-z; 4-cc;  39-d(722)            •          •

                                                        Epoxides

Ethylene  Oxide              |  l?-dd;  4-ee;  4o-mmm
Propylene Oxide            j  18-ff;  A-ee
                                                          Sufides

  bis(2-Chloroethyl)  Sulfide |              '        '
 =====BBaa===aaaaa3a=a=i;=333B333a3============aaa=a=aa3as3a==s=saa=aa33a3a3333a=3sa===3a33aa3aaaaaa3aaaaa
                                                                                              (continued)

-------
                                                                                Section No.  3.16
                                                                                Date  June 30,  1§88
                                                                                Page  18
TABLE D.   (Continued)

 Selected Compounds
                            Column Reference,  Type and Conditions,  and Kovats  Retention .Indices*
                           = = = = = ss = a = = = = = =!s = s = = = = =s = = s = = = = = = = = = cs = = = = s = = = = = = = = s = s£a» = = = sssass = = =
                                                       Halogenated
Ethylidene Chloride
Ethylene Dibromide
Ethylene Dichloride
Propylene Dichloride
1,1,1-Trichloroethane
Bromodlchlorome thane
Chlorodibromome thane
Chloroform
Carbon Tetrachloride
Dichlorodifluoromethane
Methyl Bromide
Methyl Chloride
Methylene Chloride
Tetrachloroethylene
Bromoform
Trichloroethylene
Trichlorotrifluoroethane
Vinylidene Chloride
Ethyl Chloride
Chlorobenzene
Vinyl Chloride
1 , 2-Dibromo-3-chloropropane
19-88
20-hh
21-ii
22-kk
21-ii


23-11
23-11
24-pp
25-qq
26-rr
27-88.
21-11;
19-ww
21-ii ;
21-ii;
28-zz;
29-11
19-bbb
30-ccc
37-888
39-d(565), e(427). f (604) , h(g4
4-(hhh)
19-jj; 4-1; 39-d(636), e(46o),
4-(vv)
19-JJ; 4-oo; 39-6(568), v(897)


19-mm; 4-iii; 39-d(6o6) , e(443)
19-nn; 4-oo; 39-d(665), e(502).



0), 1(994). v(896)

f (666) ,




. f(625)
f(553).




h(113D, 1(1205), v(1078)




, h(1022), 1(1090), v(1025)
h(897), i(938), v(893)



111; 32-tt; 4-oo: 39-d(5l6), h(.956), 1(1013), v(934)
33-uu; 4-vv; 39-d(8l3) , e(5?4),

34-xx; 4-1; 39-d(695), e(546),
35-yy; 4-oo
f(736).

f(665),

h(l05&), 1(1105). v(l039)

h(1009), 1(1068), v(1004)

4-jjj; 39-h(7&0). 1(792). v(738)

; 4-kkk; 39-d(842), h(1347)
, ddd; 36-eee; 4-fff
= = sa = = = = = = ss = = = = = = = sss:s = = = = = = = = s





1


The GC column references,  column types and conditions,  and  Kovats  Retention  Indices  (if  available)  are
shown in this column.   The first reference shown for  each compound is  for  the  column  and conditions
suggested for the recommended sampling method (when available),  followed by  the  column and  conditions
suggested for sampling methods with an acceptable rating.   Additional  references are  given  when
available for columns  used for analysis of the selected compound under laboratory conditions.  Some
additional references  provide Kovats Retention Indices  for  selected compounds.   For  example,  the
reference code for benzene is: 12-s, t; 9-k;  13-u;  4-p;  39-d(6s8),  e(557), 1(1039),  h(HOA),  v(9&3).
This means that for benzene the columns described in  citations  s and t in  the  List of Referenced
Columns (following Table D) were specified in the method described in  citation  12 in  Section  3.16.11;
the column described in citation k in the List of Referenced  Columns was specified in the method
described in citation  9 in Section 3.16.11;  the column  described in citation u  was specified  in the
method descibed in citation 13 in Section 3.16.11;  the  column described in citation  p was specified  in
the method described in citation U in Section 3.16.11;  and  the  columns described in.  citations  d,  e.  i,
h.and v were specified in  the method described in citation  39 in Section 3-l6.ll.  Where available,  the
Kovats Retention Indices for each of the columns under  the  conditions  given  in  their  respective
references are given in parentheses.

Note: Any column or conditions that meet the  Method 18  criteria  for peak resolution  are  considered
acceptable.

-------
                                                                Section  No.  3.16
                                                                Date  June  30,  1988
                                                                Page  19


APPENDIX I to TABLE D.

             LIST OF REFERENCED COLUMNS WITH SUGGESTED OPERATING CONDITIONS

(a).  60/80 mesh Tenax, operated isothermally at 80 C.
(b) .  Chromasorb 101, operated isothermally at 50 C.
(c).  0.2% Carbowax 1500/0.1$ SP-2100 on Carbopack C, operated isothermally at
        60°C.
(d).  20% SP-2100/0.1% Carbowax 1500 on 100/200 mesh Supelcoport, operated
        isothermally at 70°C.
(e).  Carbopak C-HT 80/100 mesh, operated isothermally at 90 C.
(f).  Porapak T 80/100 mesh, operated isothermally at l40°C.
(g).  0.2% Carbowax 1500 on 60/80 mesh Carbopack C, temperature programmed from 65°
          to 70°C.
(h).  15% tetracyanoethylated pentaerythritol on 60/80 mesh Chromasorb P-AW,
        operated isothermally at 80 C.
(i).  15% tetracyanoethylated pentaerythritol on 60/80 mesh Chromasorb P-AW,
        operated isothermally at 100 C.
(j).  10% SP-1000 on 80/100 mesh Chromasorb WHP, temperature programmed from 75 C.
(k).  20% SP-2100' on. 80/100 mesh Supelcoport, operated isothermally at 40° or
        70 C or temperature programmed from 50 C, depending on other anal-
        ytes of interest.  See referenced method for details.
(1).  10% OV-101 on Chromasorb WHP, operated isothermally at 100°C.
(m),  10% OV-101 on Chromasorb WHP, operated isothermally at 60° or 100°C.
(n).  Durapak n-octane on Porasil C, operated isothermally at 30 C.
(o).  1% SP-1000 on Carbopack B, operated isothermally at 55°C for 12 minutes.
(p) .  10% OV-101 on Chromasorb WHP, operated isothermally at 60°C.
(q) .  0.12 SP-1000 on Carbopack C, operated isothermally at 90°C.
(r).  3% OV-1 on 100/120 Gas Chrom Q, operated isothermally at 135°C.
(s).  For benzene in the presence of aliphatics, 10% 1,2,3-tris  (2-cyanoethoxy)
        propane (TCEP) on 80/100 Chromasorb P AW.
(t).  For benzene with separation of xylene isomers, 5% SP-1200/1.75% Bentone 34 on
        100/120 mesh Supelcoport, operated isothermally at 75°C.
(u).  10% OV-275 on 100/120 mesh Chromasorb W-AW, operated isothermally at 50°C or
        temperature programmed starting at 50 C for 3 minutes followed by
        15°C/min increase to 200°C.
(v).  10% FFAP on 80/100 Acid-washed Chromasorb W, operated isothermally at 125°C.
(w).  10% OV-275 on 100/120 mesh Chromasorb W-AW, operated isothermally at 50°
        or 100 C or temperature programmed starting at 50 C for 3 minutes followed
        by 15°C/min increase to 200 C.
(x).  For meta-xylene, 10% OV-101 on' Chromasorb WHP, operated isotheramlly at 60°,
        120°, or 140°C.
(y).  3% OV-225 on 80/100 mesh Supelcoport, operated isothermally at l63°C.
(z).  10% SP-2100/0.1% Carbowax 1500 on 100/120 mesh Supelcoport, temperature
        programmed from 50° to 170°C at 10°C/min.
(aa). 20% SP-2100/0.1% Carbowax 1500 on 100/120 mesh Supelcoport, operated isother-
        mally between 55° and 75°C.
(bb). Chromasorb 101, operated isothermally at l80°C.
(cc). 0.1% SP-1000 on Carbopack C, operated isothermally at 180°C.
(dd). 50/80 mesh Porapak Q, operated isothermally at 135°C.

(continued)

-------
                                                               Section No.  3-16 „
                                                               Date June 30,  1988
                                                               Page 20

APPENDIX I to TABLE D. (Continued)

(ee).  80/100 mesh Porapak QS, operated isothermally at 150°C.
(ff).  50/80 mesh Porapak Q, operated isothermally at 1^5 C.
(gg) .  10% SP-1000 on 80/100 mesh Supelcoport, operated isothermally at 50°C.
(hh).  3% OV-210 on 80/100 mesh Gas Chrom Q, operated isothermally at 50 C.
(ii).  20% SP-2100/0.1% Carbowax 1500 on 100/200 mesh Supelcoport, operated
         isothermally at 100°C.
(jj). , 10'% SP-1000 on 80/100 mesh Supelcoport, operated isothermally at 70°C.
(kk) .  3% Carbowax 1500 on 60/80 mesh Chromasorb WHP, operated isothermally at
         50°C.
(11).  1% SP-1000 on Carbopack B, operated isothermally at 120°C.
(mm).  10% SP-1000 on 80/100 mesh Supelcoport, operated isothermally at 75°C.
(nn) .  10% SP-1000 on 80/100 mesh Supelcoport, operated isothermally at 60°C.
(oo).  10% SP-1000 on 80/100 mesh Supelcoport, operated isothermally at 100°C.
(pp) .  80/100 mesh Chromasorb 102, operated isothermally at 110°C.
(qq) .  10% FFAP on 100/120 mesh Chromasorb WHP, operated isothermally at 65°C.
(rr).  80/100 mesh Chromasorb 102, operated isothermally at 100°C.
(ss).  5% OV-101 on 80/100 Chromasorb WAP, operated isothermally at 35°C-
(tt).  10% SP-1000 on 80/100 mesh Supelcoport, operated isothermally between 60°
         and 90°C.  '                   .  .
(uu) .  10% OV-101 on 100/120 mesh Supeleoport, operated isothermally at 90 C.
(w).  10% OV-101 on Chromasorb WHP, operated isothermally at 50° or 100 C.
(ww).  10% SP-1000 on 80/100 mesh Supelcoport., operated isothermally at 130°C.
(xx) .  10% OV-101 on 100/120 mesh Supelcoport, operated isothermally at 70°C.
(yy) •  50/80 mesh Porapak Q, operated isothermally at 150 C.
(zz).  100/120 mesh Durapack OPN in silanized glass, operated isothermally at 65 C.
(aaa).  10% FFAP on 100/120 mesh Chromasorb WHP, operated isothermally at 110°C.
(bbb).  10% SP-1000 on 80/100 mesh Supelcoport, operated isothermally at 105 C.
(ccc).  80/100 mesh Chromasorb 102, operated isothermally at 100°C.
(ddd) .  For sources where acetaldehyde is present, use column cited in (ccc)
         followed by a column  of  20%  GE  SF-96  on 60/80 mesh Chromasorb P AW or
         80/100 mesh Porapak T connected in series, operated isothermally at 120°C.
(eee).  10% SE-30 on 80/100 Chromasorb W, operated isothermally at 60°C.
(fff).  0.4% Carbowax on Carbopack C, operated isothermally at 50°C.
(ggg).  1.5 OV-17 plus 1.95% OV-210.
(hhh) .  5% OV-101 on Chromasorb WHP,  operated isothermally at 60 C.
(iii).  10% OV-101 on Chromasorb WHP, operated isothermally at 50° or 100°C.
(jjj).  10% OV-101 on Chromasorb WHP, operated isothermally at 100°C or 10% SP-2100
         on Supelcoport,  operated isothermally at 100 C.
(kkk).  10% SP-1000 on 80/100 mesh Supelcoport, operated isothermally at 150 C.
(111).  1% SP-1000 on 60/80 mesh Carbopack,  temperature programmed- starting at 40 C
         for 3 minutes,  followed by 8 C/min increase to 200 C.
(mmm) .  15% FFAP on Anakrom A.                                 .
(nnn) .  10% FFAP on 80/100 mesh Chromosorb W AW-DMCS, operated isothermally at 52°C.

-------
                                                                              Section  No.  3-16
                                                                              Date June 30.  1988
                                                                              Page 21
        TABLE E.   RECOMMENDED CALIBRATION  TECHNIQUES FOR SELECTED ORGANIC COMPOUNDS BY METHOD 18
  ===============================================================================================
                                  Methods  for  Direct  Interface
                                     and Tedlar  Bag Samples
Selected Compounds
===========================
                                 Gas
                              Cylinders
                                              Gas      |   Liquid
                                            Injection   | Injection
                                              into     |    into
                                            Tedlar Bag  | Tedlar Bag
                                                                        Methods for Adsorption
                                                                             Tube Samples
                           Prepare    |   Prepare
                         Standard  in  |   Standard
                         Desorption   |   on Tube
                           Liquid     |  and Desorb
                                 ================================================================
                                             Alcohols
Methanol i
Ethanol
Isopropyl Alcohol
n-Propyl Alcohol
n-Butyl Alcohol
T-4
U
U
u
, u
N •
N
N
N
N
U
U
U
' U
U
R-6 | T
R-7 | T
R-7 | T
R-8 | T
R-8 | T
                                             Alkanes
Cyclohexane
Hexane
                                 T-4
                                 T-4
                             R-9
                             R-9

                                             Alkenes
Ethylene
Propylene
                                 T-4
                                 T-4
                                              Dienes
1,3-Butadiene
Hexachlorocyclopentadlene
                                 A-10
                                  U
 IN       |
1===========
   Aromatics
R-10
 N
                                                       ==============
 R-41
 R-ll
Benzene
Mesi tylene
Ethylbenzene
Cumene
Xylene (m- , o- , p- )
Toluene
Styrene
2-Napthylamine
|R-12(SRM 1806)
1 • "
. . "I "
1 u
1 T-4
| T-4
1 • "
1 "
• N'
N .
N
N
N
N
N
U
A-12
U
U
U
u
u
u
u
R-9. 13
u
R-13
R-13
R-13
R-9. 13
R-13
R-14
T
U
T
T
T
T
T
T
====================================================
                                             Ketones
                                                   ===============================
Acetone
Methyl Ethyl Ketone
Methyl Isobutyl Ketone
                                  U
                                 T-4
                                 T-4
R-15
R-16
R-15
                                     |      T
                                     I      T
                                     |      T
                                         ===============
                                             Cpoxides
                                                       =========================================
Ethylene Oxide
Propylene Oxide
                                 T-4
                                 T-4
                            R-17
                            R-18
=============================================
                                             ==================================
                                             Sulfides
                                                                              ==================
bis(2-Chloroethyl)  Sulfide |
                                                                            u      I      u

                                                                                    (continued)

-------
                                                                               Section No. 3.1§
                                                                               Date  June  30,  1988
                                                                               Page  22
TABLE E.  (Continued)
Selected Compounds
Methods for Direct Interface
and Tedlar Bag Samples
Gas
Cylinders
Gas
Injection
into
Tedlar Bag
Liquid
Injection
into
Tedlar Bag
Methods for Adsorption
Tube Samples
Prepare | Prepare
Standard in ( Standard
Desorption | on Tube
Liquid 1 and Desorb
                                           Halogenated
Ethylidene Chloride
Ethylene Dibromide
Ethylene Dichloride
Propylene Dichloride
1,1, 1-Trichloroe thane
Bromodichlorome thane
Chlorodibrornome thane
Chloroform
Carbon Tetrachloride
Dichlorodifluoromethane
Methyl Bromide
Methyl Chloride
Methylene Chloride
Tetrachloroethylene
Bromoform
Trichloroethylene
Trichlorotrifluoroethane
Vinylidene Chloride
Ethyl Chloride
Chlorobenzene
Vinyl Chloride
1 ,2-Dibromo-3-chloropropane
U
T-4
R-21
T-4
R-21
U
U
R-23
R-23
U
U
U
R-21
R-2KSRM 1809)
U
R-21
R-21
T-4
U
T-4
R-30
U
N
N
N
N
N
U
U
N
N
U
U
U
N
N
N
N
N
N
N
N
A-30
N
U
N-31
A-21
U
A-21
U
U
A-23
A-23
N
N
N
A-21
A-21
U
A-21
A-21
U
U
U
N
U
R-19
R-20
R-19
R-22
R-19
U
U
R-19
R-19
R-24
R-25
R-26
R-32
R-33
R-19
R-34
R-35
R-28
R-29
R-19
R-3&
R-37
T
T
T
T
T
U
U
T
T
T
T
T
T
T
T
T
T
T
T
T
T
T
Rating Code
 R = Recommended.
 A = Acceptable.
 T = Theoretical.
Based on actual source test experience (sampling and analysis)  this  method
is valid and is the method of choice among Method 18 users.

Based on actual source tests or similar source test experience  (sampling and
analysis), this method is valid.   The tester must evaluate  for  specific
test.

Method has no documented sampling and analysis experience,  but  in  theory
could be valid.
 N = Not Recommended.  Based on actual source tests  or  similar  source  test  experience   and/or
                      theory,  this method is invalid.
 U = Unknown.
                      Method has no documented experience  and  the  theoretical  aspects  are not
                      conclusive.   The tester must  demonstrate  that  this  calibration method  is
                      valid.
   The rating codes for calibration procedures  are  based  on  procedures  specified  in  applicable
sampling and/or analytical methods.  For example,  the  rating code  for benzene  is:  R-12(SRM  1806);
N; A-12; R-9,13;  T.  This means that for benzene,  the  recommended  calibration  procedure  for  direct
interface and Tedlar bag samples involves the use  of gas  cylinders  with the  procedures described
in citation 12 in Section 3.16.11 and Standard  Reference  Material  1806  (available  from the
National Bureau of Standards.  Gaithersburg,  MD);  calibration standards  for benzene prepared  by gas
injection into Tedlar bags is  not recommended;  calibration standards prepared  by  liquid  injection
into Tedlar bags  is acceptable following the procedures described  in citation  12  in  Section
3.16.11; preparation of calibration standards in  desorption  liquid  is the recommended procedure
for the adsorption tube methods described in citations 9  and 13  in  Section 3-l6.ll;  preparation of
calibration standards on adsorption tubes followed  by  desorption is  theoretically  valid  for  use
with adsorption tube samples.

-------
                                                                Section No. 3.16.1
                                                                Date  June  30,  1988
                                                                Page 1
1.0  PROCUREMENT  OF  APPARATUS  AND  SUPPLIES
        For Method 18,  a number of different sampling and analytical'procedures are
 considered acceptable for  the identification and  measurement of  the  majority of
 gaseous organic compounds emitted from industrial  sources.   Persons attempting to
 apply these procedures are  advised  to consult the tables presented in the Method
 Highlights Section.  The Method Highlights Section is intended to provide guidance,
 based on current EPA methodology, for selection  of the most suitable sampling and
 analytical protocols for  organic  compounds of  interest  to Federal, State, and local
 agencies.   For  situations  where EPA  methodology is not applicable,  guidance for
 selection  of  sampling  and  analytical  protocols based on  methodology  from other
 reputable  organizations  is provided.    Once  a  suitable  sampling  and analytical
 protocol has  been selected,  then procurement of the necessary  apparatus  and sup-
 plies can  begin.
        A number of  the sampling  and  analytical  methodologies covered by Method 18
 are common to  both presurvey sampling  and analysis  and  final  sampling and analysis.
 Presurvey  sample collection can  employ either glass  sampling flasks (not employed
 for final  sampling), Tedlar bags, or adsorption tubes.   Apparatus to determine the
 moisture content, temperature, and static pressure of  the  source emissions may be
 required during presurvey sampling if this information is not available from plant
 personnel.  Method  18 also  describes  several different calibration techniques for
 use  depending  on  the  available  calibration  materials   and  the sampling  and
 analytical techniques used.  Confirmation of  target compounds in presurvey  samples
 may require analysis  by means other  than  GC alone, such  as GC/mass  spectrometry
 (GC/MS) or GC/infrared spectrometry  (GC/IR).  For  the  final sampling,  in addition
 to Tedlar bags and  adsorption tubes, direct interface  sampling and  dilution inter-
 face sampling  are described.  Analysis of the  final samples utilizes the procedures
 developed  and  optimized during presurvey  sample  analysis.
        The descriptions  of  the  apparatus and supplies  that  follow apply to items
 needed  for both  presurvey  and final  sampling and  analysis, except as noted.   As
 described  above,  all of the following equipment  may not be required.  The following
 procedures and descriptions are only  provided as  guidance to  the  tester and  are not
 requirements  of  the method.   Table  1.1 at  the end  of  this section contains   a
 summary of quality assurance activities for  procurement and acceptance  of apparatus
 and supplies.

 1.1 Sampling

        Guidance for  the  selection of  a suitable  sampling technique  for  a particular
 compound can be found in Table B  of  the Method Highlights Section.

 1.1.1  All Sampling Procedures - The  following  apparatus will be required for all
 presurvey  and  final sampling procedures..   Use  of  alternative  equipment requires
 the approval of the  Administrator.

        Sampling System Check - Because of the number of sampling systems, volatile
 organic compounds,  and process operating conditions,  the exact criteria for check-
 ing the sampling system  can only be  determined  using the presurvey sampling data.
 Upon receipt  of  all the components  to  construct the  sampling system,  the system
 should be  assembled  and  checked  over the intended  range of  use (i.e.,  sample flow
 rate,  duct temperatures).

-------
                                                                Section No. 3.16^.1
                                                                Date  June  30,  1988
                                                                Page 2

     .  Sampling Probe  -  The sampling probe should  (1)  be constructed of stainless
steel, Pyrex glass,  or Teflon tubing, (2) exhibit  an  outside  diameter (OD)  of 6.4
mm, (3) be enlarged at the duct end to contain a glass wool plug, and  (4) possess a
heating system  capable . of  maintaining  the sample  temperature at 0°  to  3°C above
duct temperature.  The expanded section of the probe must be packed with glass wool
prior to sampling.  The probe outlet must have a fitting suitable for attachment to
the sample  line.   A probe  approximately  1.1  m (4  ft)  long  is usually sufficient;
the exact  length can  be  determined during the  preliminary  survey.   The selected
probe  material  should be  nonreactive toward with  the sample gas  constituents  so
that it will  not bias the  analysis,  as  well  as appropriate to  withstand the duct
temperature...•".                                                 .
       Upon  receiving   a  new  probe,  visually  check  it  for   adherence  to
specifications  (i.e.,  the  length  and  composition  ordered).    Check  for breaks,
cracks; and leaks.   Leak  check  the probe  and  check the probe heating system during
the.sampling system  check described  above.   The probe should be able to maintain
the required temperature at the desired flow rate and remain leak free.

       .Sample Line and Connecting Tubing - The sample  line  is generally 6.4-mm OD
Teflon tubing.   Sample lines will  require heat-tracing to prevent condensation of
sample constituents  during sampling at  some sources.   The sample temperature must
be  maintained  at 0°C.  to  -3°C above  the  source temperature.    The capacity  of the
heating system should  be  sufficient, for its intended use.   Upon receipt or during
the system check, the  sample  line should  be checked to ensure  that it is leak free
and will maintain . the desired temperature at  the  desired flow  rate.  It  should be
noted  that  heat-traced  sample  lines require  a significant amount  of electrical
current to maintain the higher, temperature levels.   The electrical requirements and
the weight of the heat-traced line, .should be  taken into, account when designing the
sample train.           .    .

       Quick Connects  -  For connections  on the  sample lines,  gas  sampling valve,
the pump unit, .cylinders,'  sample  bags, and calibration gas bags,  quick connects or
the equivalent  are  needed. .  When the connects  come into contact with the  sample
gas, they  should be constructed  of stainless steel.   It is  also useful  to  have
self-sealing quick connects  on  the sampling  bags.   The quick  connects can be leak
checked during.the system check.

       Barometer  -  A  mercury,  aneroid,   or  other  barometer capable  of  measuring
atmospheric pressure to  within-2.5 mm (0..1 in.) Hg may be  used;  however,  in many
eases  the>  absolute -barometric  pressure  can  be  obtained  from  a nearby  weather
service station.., If-the elevation of the sampling point is higher than that of the
weather station, .the  reported barometric pressure  is  reduced  at a rate  of  2.5  mm
Hg/30 m (0.1 in.  Hg/100 ft) of elevation difference; if the sampling point is lower
than the weather  station,  the pressure  is increased at the  same  rate.   Note:   The
barometric  pressure from the .weather service station should not be corrected to sea
level.
  .-,  .   Check the field  barometer  against- a  mercury-in-glass  barometer (or  its
equivalent).   If the field barometer  cannot be adjusted to agree with the mercury-
in-glass -barometer,  it is not acceptable and should be repaired or replaced.

       Moisture Determination -  A moisture determination  may  be  required.   Two
techniques  can  generally be  ..used:  (1)  Method   4  or  .(2)   wet  bulb/dry  bulb
thermometers.   If Method 4 is used,  the  tester should  refer  to Section 3-3 of this
Handbook.   If the wet bulb/dry bulb thermometers  are used, both thermometers should

-------
                                                                Section No. 3.16.1
                                                                Date June  30,  1988
                                                                Page 3

be  accurate  to  within  1°C.    Upon receipt,  the thermometers  should be checked
against a mercury-in-^glass thermometer to ensure that they are reading properly.

       Flow Rate Determination  - The flow rate  in the duct may have  to  be deter-
mined  for  some emission standards.   If the  flow rate  is  to be  determined,  the
tester should  refer  to Section  3-1 of this Handbook  and  meet  the  requirements and
follow the procedures of this method.

1.1.2  Glass Sampling Flask Sampling Technique -  The following  apparatus and rea-
gents will be  required for the .collection of  samples (presurvey only)  using glass
sampling  flasks.     Use  of  alternative  equipment requires  the  approval of  the
Administrator.

         Purged  or  Evacuated  Glass Sampling  Flasks  -  Presurvey  samples can be
collected  in  precleaned double  ended  glass  sampling  flasks  possessing minimum
capacities of  250  ml.   Teflon stopcocks,  without   grease,  are preferred.   Upon
receipt,  flasks should be  checked  to ensure  that they are not broken.  Flasks must
be cleaned prior to  use.  The cleaning procedures  are described later in Subsection
3.2.  If the flasks  do not meet these requirements, replace or reclean.

1.1.3  Tedlar  Bag/Evacuated Container  and  Adsorption Tube Sampling - The  following
apparatus will be  required for the  collection of  presurvey  or final samples using
adsorption tubes or Tedlar bags housed  in evacuable  containers.   If the  apparatus
are  purchased  separately,  each item should be  checked  individually  as   described
below.  Following this,  all components should be assembled, as they will be used in
the  field and  then checked using the following procedures:
       1.   Assemble the sample train as described in Subsection 4.3-1•
       2.   Leak check the  train as  described in Subsection 4.3-1-
       3.   Attach  a primary gas test  meter  to the inlet of the sample train and
            pull the desired  flow  rate through the sample train for the typical
            sample run time.  The  measured volume should be within 10% of the
            calculated volume or rate.   If the system does not meet these
            requirements, replace or repair and then  recalibrate.

       Tedlar  Bags  (For  Sampling  and to  Prepare Gaseous  Calibration Standards)-
Bags used  to collect field samples  and  prepare  gaseous calibration standards must
be constructed of a  suitable  material, be  leak free,  and have the proper  fittings.
Typically, self-sealing  quick disconnects  are  used on  the sample  bags.   Tedlar is
the  material of  choice  for the sample  bags,  however other materials  may be used
successfully.  If the  sample bags are constructed  by  the  tester, they are  generally
double-sealed.   The  exact  bags  to  be used  in the field test or for making calibra-
tion standards must  pass three criteria as follows:
       1.   Bags must pass the leak  check as described in Subsection 4.3-1-
       2.   The organic  components  that  are  to be collected in  the bags  should be
            placed in a  bag at about  the  same concentration  for  which it will be
            used, and the organic  concentration  in  the bag determined as soon as
            possible  after  this.  The  organics should then remain  in the  bag for a
            period   equal   to  the  time  anticipated  between  field  sampling  and
            analysis.   The concentration,  upon  reanalysis,  must be  within  10% of
            the original concentration.
       3.   Next, the bag should be emptied and refilled with zero air or nitrogen.
            It  should  be  allowed  to   sit  for  at  least  2  hours  and  then  be
            reanalyzed;  the concentration of  the organic(s)  in  the bag must be

-------
                                                                Section No. 3-16..1
                                                                Date  June  30,  1988
                                                                Page 4

            less than 10% of the original concentration.
If the bags do not pass the leak check, they must be reconstructed.  If they do not
pass  the second  and third  criteria,  a  different material  of construction  or a
different sampling technique must  be used.   If the bags do not meet these require-
ments, replace, use an alternative technique or use a different sampling technique.

       Rigid  Leak-Proof Container  -  Rigid  leak-proof containers  must be  of the
proper size to fit the  bags  and are generally made of rigid plastic.  However, the
material of  construction  is typically not  important since the  container  does not
come  into  contact with the  sample gas.   Containers usually  have a  clear  top or
window in them to check that  the bag does not overfill during testing.  The top of
the container must have connections to attach the  sample  probe to the outside and
the  sample  bag  to  the inside.    A  connection for the sample  pump  must  also be
available  on  the container.    Upon  receipt  or  construction  of  the  container,
assemble the system  as  it  will be used in the field  and then  leak check it at the
maximum vacuum anticipated.   Inflate a bag to the  degree  it  will  be filled in the
field  and  check  that  the  bag  can be removed  after it has  been  filled  to allow
external heating.   If  the bag system is designed  to keep the bag at a specified
temperature,  then  the heating  system   must  be  checked  as  described  below  in
Subsection  1.1.6.    If the  container does  not meet these  requirements,  modify,
repair, or replace it.

       Pump -  For the indirect sampling  technique  (pump after the bag or charcoal
tube) , any  pump  of proper capacity can be used.   If the pump  is  to be used for a
direct sampling  technique (pump  in the  sample line),  the pump internals  must be
leakless and made of stainless steel or, preferably, Teflon.   Upon receipt, check
for proper specifications.  If the pump does not meet the specifications, repair or
replace it.

       Flowmeter  -  The flowmeter must be  of  the  proper  flow  rate  range.   Upon
receipt,  check  the specifications  and then  calibrate  as   described  in Subsection
2.1.3.  If the flowmeter does not meet the requirements, replace or recalibrate it.

       Adsorption Tube  - An  absorption tube must  (1) be of adequate capacity, (2)
contain the proper adsorption  material, and  (3) consist of a primary and secondary
section.   The  selection of the  proper  type  and size of adsorption tube  should be
based on previous experience  (including  the  literature and tables  in the   Method
Highlights)   or laboratory  evaluation.    The  selection and/or  evaluation  of  the
proper adsorption tube is  described in  detail  in Subsection 3-4.   The  criteria
shown in Subsection 3-4 must be met or the tubes must be replaced or modified.

       Personnel  Sampling Pump  - A  personnel  sampling  pump can  be  used  for
collecting  adsorbent  tube samples.   It  must sample  at or  be adjustable  to the
proper flow  rate  range.  Upon  receipt,  check the  specifications  and calibrate as
described in Subsection 2.1.4.   If it  does not meet the specifications, replace or
calibrate it.

1.1.4  Direct  Pump Sampling  Procedure - The  direct pump  sampling procedure  will
require the same  apparatus described  in Section 1.1.3  for  bag and adsorption  tube
sampling.  The  only  difference is  that the  pump  internals must be constructed of
materials that will not interfere in the analysis  (i.e., Teflon or stainless steel)
and  the  rigid container  does not  have  to  be leak  free.    The  system should  be
assembled,  leak checked and then  the flow rate  checked  as  described above.   If the

-------
                                                                Section No. 3.16.1
                                                                Date June  30,  1988
                                                                Page 5

system does  not meet the criteria,  then  it must be replaced  or  repaired  and then
leak checked and recalibrated.

1.1.5  Explosion Eisk Bag Sampling Procedure - The following apparatus, in addition
to the apparatus described  in Subsection  1.1.3,  will be required for collection of
bag samples in areas where  there is any possibility of explosion.  These procedures
assume that the sample gas  collected is not above the lower explosive limit.  If it
is,  a complete  safety plan  should be  developed and  reviewed  by  the plant  and
tester.  This Handbook will not attempt  to describe  the procedures used to collect
explosive gases.
  The  major  concern in most  areas having  an  explosion potential is  that no open
flames or non-intrinsically safe electrical equipment be used.  The first approach
to sampling  in  these areas  is  to  remove  the electrial  systems (i.e.,  pumps)  to an
area  that  is not explosive.   This can be accomplished by  running the vacuum line
from  an  area  that  is not  explosive  to  the  sample bag or  charcoal  tube  in  the
explosive area.  The vacuum line can be run up to 200 to 300 feet with no problems.
Sampling is then conducted  in the normal manner.
       Another approach described in Method 18 uses a steel canister to provide the
vacuum source.   This approach  is  difficult and can still  be  hazardous  because it
involves handling  a steel  container,  and possibly a stainless steel  probe,  in an
explosive area.  Another alternative is the use of an intrinsically safe pump, such
as  a  personnel  sampling pump, with  adsorption  tubes  or Tedlar bags housed in
evacuable  containers.    Any   system  that  is  purchased  or   constructed  must  be
leakless,  be able  to  control  the  flow  rate  properly,  and meet  all  plant safety
requirements.

1.1.6   Heated Bag  Sampling Procedure  - This  procedure must  be  used  in the event
that  condensation  is observed  in  the  bag and/or sample line  during testing or if
the sample bag retains more than 10% of the sample concentration based on the post-
test  sample  bag retention check.  The apparatus  described  below  will  be required,
in addition  to  the  apparatus  described in  Subsection 1.1.3-   Both the sample line
and container must  be  heated  to maintain  the bag at a specified temperature (i.e.,
0°C to 3°C above source temperature).  The sampling system is checked in the manner
described  above,  except  that  the  heating system must  also  be  considered;  check
procedures should consider  use  of the  system at ambient temperatures less than the
laboratory temperature  (including wind chill  factors).   The entire  surface of the
sample probe  and the sample  bag must  be maintained at  the specified  temperature.
A possible alternative to maintaining the bag at  the specified temperature is the
addition of  external  heating  with heat lamps  prior  to  analysis.   The  exact system
that will be  suitable for any  given source should be determined  prior to testing,
if possible.  The operation and checks of heated sampling systems are  described in
Subsection 4.3-   If  the  system does  not meet all  the criteria, use a different
approach or repair  the system and recalibrate.

       Heated  Bag   Sample  Container - The  heated  bag sample  container  must  be
capable  of  maintaining  the  entire  bag  at   the  specified  temperature.    If  an
electrical source is  used to heat  the container,  the tester must be  aware  of  the
additional explosion  potential that is created.   One  check  on the system  can be
made with a  thermocouple  in the sample cavity; this check  of  the system will  not,
however,  demonstrate  that  all  the  surfaces   are maintained   at  the  required
temperature.   All external  surfaces of the container should be well insulated.   A
visual check of  the system  should reveal if the  system  appears  to be  sufficiently
insulated.    If the  system  allows  the  bag to have  cooler surfaces,  the  posttest

-------
                                                                Section No. 3.l6vl
                                                                Date June  30,  1988
                                                                Page 6
retention check on the bag will probably fail.
       Heated Sample Lines and  Probe  -  The sample probe and  sample  line must also
be capable of maintaining the specified temperature.   They should also not have any
cooler surfaces.   These can  be checked by pulling  the desired flow  rate through
them  and  then  checking  the   temperature  in  the  line  and/or  probe  with  a
thermocouple.    Proper  insulation   is  necessary  to maintain  the  temperature
throughout the entire length.   If  the temperature  cannot be  maintained,  replace or
repair the-line and/or probe and then recheck.

1.1.7  Prefilled Tedlar  Bag  Sampling Procedure - This  procedure is  an alternative
to the heated bag sampling  procedure.   The  following apparatus and  reagents are
required for this procedure,  in addition to the apparatus  described  in Subsections
1.1.3  and 1.1.4.    The  prefilled bag sampling  system  is  used   to  dilute  the
concentration of the condensibles  below saturation.   This system can  also be used
to dilute  the  gases  to below  the lower  explosive  limit.   The major  difference
between the  prefilled  bag sampling  system and the  other  bag sampling  systems is
that  the  volume of  diluent  gas  added to  the sample  bag and  the  volume of gas
sampled must  be accurately measured.   The dilution must  be accounted  for in the
calculation of the measured gas  concentration.  Therefore, the diluent gas must be
added with  a  calibrated dry gas  meter or a  calibrated  flowmeter;  then,  during
sample collection,  the gas collected  must  be  accurately measured using a flowmeter
or a  metered pump.   To  obtain  the required  accuracy,  the flowmeter  and  pump are
placed in the  sample line  prior to the sample  bag.  Since condensation may occur,
the flowmeter  and pump must  be  housed in a heated box.  This system is checked in
the same  manner as a dilution  system (see Subsection  1.1.9).   The check for the
prefilled system is described in Subsection 4.3-4.

       Heated Flowmeter  -  A  calibrated heated  flowmeter is  required to accurately
determine the  volume of  gas  sampled.   The flowmeter should  be housed  in  a heated
box  that  will maintain  the  specified  temperature.    The  flowmeter  should  be
calibrated as  described  above in Subsection  2.1.3-   If the  criteria  are  not met,
replace or  repair and  then  recalibrate.   A metering type pump may be  used to
replace the flow rate meter and the pump.

      'Positive Displacement Teflon-Lined Pump - A positive displacement pump lined
with Teflon or  constructed of stainless  steel,  of proper  capacity  and contained in
a heated  box is required.   A Teflon-coated diaphram-type pump  that can withstand
120°C and  delivers  1.5  liters/minute is typically used.   Upon  receipt,  check the
pump  for  capacity  and then  conduct  a leak check on  the  pump.   The  pump must be
leak  free  at  all  vacuum settings.  The heating system will  be  checked  during the
sampling system check.   If  the pump  is not  of the  correct   capacity  and  not leak
free, then replace or repair it.

       Heated Box for Flowmeter and Pump -  The flowmeter and  pump must be contained
in a  heated  box to maintain  the proper temperature.  Construct the box such that
the  temperature  can  be  controlled  and monitored.   After construction, check the
system to ensure that it will maintain  the desired temperature(s).   If it  will not
maintain the temperature(s),  repair the unit.

1.1.8' Direct  Interface  Sampling Procedure -  A heated probe, heated  sample  line,
heated gas  sampling  valve, needle  valve,  and  charcoal adsorber are  required for
direct interface sampling.   The required apparatus  and reagents pertaining to the

-------
                                                                Section No.  3.16.1
                                                                Date June 30,  1988
                                                                Page 7

gas chromatograph  are described  below in  Subsection 1.2.   After  the  individual
components are  checked as shown below, the sampling system should  be  assembed as
shown in Subsection 4.3-6 and checked using the following procedures:
       1.   Turn on  the  heating system  and adjust  to  the maximum  temperature at
            which it is to be used.
       2.   Connect the inlet to the  sampling  probe.  After the heating system has
            stabilized at  the  temperature  setting,  turn on the pump  and  evacuate
            to about  10 in.  of  H20.   The system must be leak  free;  no flow should
            be observed from the charcoal  adsorber  system.   If the  system is not
            leak free, repair the system.
       3.   After  the system has  been shown  to  be  leak  free, adjust  the needle
            valve until the  flow  rate that will be used in  the field is obtained.
            The temperature  at  the discharge of the  heated  sample  valve should be
            at  the  set temperature.  Insert a thermocouple  into  the inlet  of the
            probe to  insure  that  the  first several feet of  the probe and  line are
            properly  heated.   If the  set  temperature is not  obtained,  repair the
            system  or use  the  temperature obtained for a  recalibration   of the
            temperature setting.

       Heated Probe  and  Sample Lines  -  The sample  probe and  sample  line  must be
equipped with a heating system  and insulation.  All  of  the interior surfaces must
be maintained at  the temperature setting.   Although  all the interior surfaces can
not be easily checked, installing  proper  insulation and  following the system check
shown above should be  sufficient to determine  the  adequacy of the probe and sample
line heating system.

       Heated Gas  Sampling  Valve  -  A heated sampling  valve  (which  includes the
sample  loops)  is  required  to  maintain the  sample  injected  into  the  GC  at the
desired  temperature.   The sample valve and loop are  generally enclosed in  an oven
in which the temperature  can be controlled and monitored.   Upon receipt, check the
temperature controller.

       Charcoal Adsorber - The charcoal adsorber is required to remove the organics
from the excess flow  through the  system.   Since the charcoal adsorber is used only
for tester safety,  there  are no requirements on the  adsorber.   However, since the
charcoal  will  be  spent  with  time,   the   tester  should  change  it  periodically.
Alternatively, the flow can be vented at a  safe distance away from any personnel.

1.1.9   Dilution Interface Sampling Procedure  - In addition  to the apparatus des-
cribed  in  Subsection 1.1.8, dilution  pumps,  flowmeters  and valves  which  are con-
tained in  a heated box,  and diluent gases  are  required  for the dilution interface
system.   The  calibration  of the dilution  system  is described in  Subsection 2.2.
The individual  components  should  be  checked  as  shown  below  and  then  the system
should be  calibrated  as  described in Subsection 2.2.  If the system does  not meet
the calibration requirements, it should be  replaced, or  repaired and recalibrated.

       Dilution Pumps -  Two Model A-150  Komhyr Teflon  positive  displacement-type
pumps,  or  equivalent  models capable  of  delivering  150 cc/minute,  are required.
Alternatively, calibrated  flowmeters  can be used  in conjunction with Teflon-coated
diaphram  pumps.   Upon  receipt  calibrate  the pumps   or  flowmeter  and   pump  as
described  in   Subsection  2.1.    If  the  pumps   do not  meet   the  calibration
requirements,  replace or repair and then recalibrate.

-------
                                                                Section No. 3.16.1
                                                                Date  June 30, 1988
                                                                Page  8

       Flowmeters  -  Two flowmeters are  required  to measure the  diluent  gas, at a
rate of about 1350 cc/minute.  Upon receipt, the flowmeters should be calibrated as
described  in Subsection  2.1.    If the  flowmeters  do  not  meet  the  calibration
critera, replace or repair and then recalibrate.


       Diluent Gas  -  Diluent gas in cylinders  fitted with regulators are required
for sample dilution.  Nitrogen or hydrocarbon-free air can be used depending  on the
nature of  the source  gases.   Alternatively,  ambient air  can  be cleaned and dried
with charcoal and  silica  gel.  The organics in the  dilutent gas must be below the
detection limit.

       Heated Box for Sample Dilution System - The pumps and control  valves must be
housed in a heated box  to  control and  monitor the temperature.  After construction
or receipt, check  the temperature control system.   If the box cannot maintain the
desired temperature, replace or repair and recheck.

1.2  Sample Analysis

       The analysis  of  Method 18 samples  requires the use of  a gas chromatograph
(GC)  regardless  of  the technique  used  for  either  presurvey  or  final  sampling.
Guidance for the selection of suitable GC detectors  is provided in Table C in the
Method Highlights Section.   As a starting point  for the analysis of the presurvey
sample,  Table  D  in  the  Method  Highlights  Section provides  guidance for  the
selection of a suitable packed GC column.  Any interferences  with the GC analysis
may be source-specific, so  the most suitable  analytical  system must be established
using the presurvey samples.   The following apparatus will be  required for the GC
analysis.

1.2.1  Gas Chromatograph  - A GC equipped with  a  suitable  detector as specified in
Table  C  in  the Method Highlights  Section.    The  GC shall  be  equipped with  a
temperature-controlled  sample  loop  and valve  assembly for  analysis  of  gas samples
or  a  temperature-controlled  injection port  for  analysis  of  liquid samples from
adsorption tubes.   Use  of alternative techniques  for introducing samples into the
GC requires the  approval  of  the  Administrator.   The GC should  be  equipped  with a
temperature-controlled  oven,   while  a  temperature-programmable oven may also  be
required for some  analyzers.  Method  18 may be  used to  quantify  gaseous organic
compounds  at  concentrations  ranging from  about  1  part-per-million (ppm) to  the
upper  range  governed  by  detector  saturation  or column  overloading.   For  the
combination of GC  options chosen,  the  lower  limit of quantitation,  as defined by
Knoll'12 ,  for the  target organic  compounds  should be less  than  the emission  limit
for the particular source being tested.

1.2.2   GC Column  - Guidance  for the  selection  of  the  appropriate GC  column is
provided in  Table  D  in  the  Method  Highlights  Section.   The columns  listed  in
Appendix I to  Table D  have  been found  to  work for  analysis of the corresponding
organic compounds  under certain  conditions.    Since interfering compounds may be
source-specific,  Method 18 permits the use of any GC column, provided the following
precision and accuracy are achieved:

    Precision:  Duplicate analyses within 5 percent of their mean value.
    Accuracy:   Analysis results of  an  audit sample within  10 percent of the
               prepared value.

-------
                                                                Section No.  3.16.1
                                                                Date June 30,  1988
                                                                Page 9


In addition,  resolution  of interfering  compounds  from target compounds  should  be
achieved.  For determining whether  acceptable  resolution  has  been achieved,  follow
the procedures described in Appendix  C  "Quality  Assurance Procedures"',  Procedure 1
"Determination of Adequate Chromatographic Peak Resolution."'*3

1.2.3   Recorder -  A linear  strip  chart  recorder  is required,  as a  minimum,  to
record  the  GC detector  output.    Alternatively,  an  electronic  integrator may  be
used and is generally recommended.

1.2.4   Recorder  or Electronic  Integrator Paper  -  Consult  operator's  manual  or
manufacturer for correct type.

1.2.5   Regulators - Gas cylinder regulators  will be  required  for use of  the gas
cylinders described in Subsections 1.3-1, 1.3-2,  and 1.3-3-  Consult with suppliers
of gas cylinders to determine the proper type of regulator required.

1.2.6   Tubing and Fittings - Tubing  and fittings will be required to connect the
gas cylinder regulators to the GC.

1.3  Reagents and Glassware

       The exact  reagents  and glassware required depend  on  the  sampling procedure
chosen,  the  calibration  techniques  to be used,  and  the particular requirements of
the GC system.

1.3.1   GC Carrier  Gas  -  The carrier gas  selected must  be hydrocarbon-free.   The
type of  carrier  gas depends on  the type of GC detector  and  GC  column being used.
Consult   the GC  operator's  manual,  the GC  manufacturer,  and/or  the  column
manufacturer  for recommendations  on the optimum carrier gas for a particular appli-
cation.

1.3.2   Auxiliary  GC Gases - Certain  GC  detectors  will require  auxiliary gases for
proper  operation.    Consult  the  GC  operator's manual or the GC  manufacturer for
recommendations on  a particular  application.

1.3-3    Calibration Gases   -   These  include  cylinder   gases  containing  known
concentrations  of  target organic  compounds  for  preparation   of GC  calibration
standards,  direct use as  GC  calibration  standards,  or  calibration of a dilution
interface system.   If gases are not available in the required concentrations for GC
calibration,  procure the  reagents and  glassware described  in  Subsections  1.3*4
through  1.3.7.

1.3-4   Zero  Gas  -  Hydrocarbon-free  air or nitrogen, for preparing  gaseous  cali-
bration  standards from calibration gas cylinders or liquid organic compounds.

1.3.5  Liquid Organic Compounds - Pure or high purity liquid  (occasionally gaseous)
samples of all the organics for which calibration standards will be prepared.

1.3.6   Syringes  - Calibrated, gas  tight 500-, 10-,  and  1.0-microliter sizes  with
maximum  accuracy,  for  preparing  gaseous calibration   standards,  for  preparing
adsorption tube standards, and  for injection of liquid standards  and  samples  into
the GC.  Other size gas tight syringes may be appropriate.

-------
                                                                Section No. 3.16.4
                                                                Date June  30,  1988
                                                                Page 10


1.3-7  Midget Impinger/Hot Plate Assembly  -  To  prepare gaseous standards in Tedlar
bags  from liquid  organic compounds  requires  a  midget impinger  equipped with  a
septum and a tee on  the  inlet  stem  and a boiling  water bath on a hot plate.  A dry
gas meter, previously described in Subsection 1.1.1, is also required.

1.3-8 . Screw  Top Septum  Vials  -  For preparation  of adsorption  tube standards and
samples,   J-ml amber screw top septum vials with Teflon-lined septa are required.

1.3.9   Desorption Liquid  -  For  preparation  of  adsorption  tube  standards  and
samples,  desorption liquid is  required.  For the correct  desorption liquid,  refer
to the appropriate NIOSH method for the target compound(s)  referenced in Table B in
the Method Highlights Section.

-------
                                                                Section No.  3-16.1
                                                                Date June 30,  1988
                                                                Page 11

      Table 1.1.   ACTIVITY MATRIX FOR PROCUREMENT OF APPARATUS AND SUPPLIES

Apparatus
All Sampling
Procedures
Sampling system
check
Sampling probe
Sample line and
connecting tubing
Quick connects
Barometer
Moisture
determination
Flow rate
determination
Glass Sampling
Flask Technique
Purged or evacu-
ated sampling
flasks

Acceptance limits
Maintain proper
flow rate and
temperature
Proper material
of construction
and capable of
maintaining proper
temperature
Constructed of
Teflon and capable
of maintaining
proper temperature
Stainless steel
construction and
leak free
Capable of measuring
atmospheric pressure
to within 2.5 mm
(0.1 in.) Hg
See Section 3-3 of
this Handbook
See Section 3.1 of
this Handbook
Double ended glass
flask with Teflon
stopcocks

Frequency and method
of measurement
Upon receipt, conduct
check in specified
subsection
Visually check and
then run heating
system checkout
Visually check and
then run heating
system checkout
Visually check and
conduct leak check
Check against mercury
in-glass barometer or
equivalent
(Sec. 3-5.2)
Same as in Section 3 • 3
Same as in Section 3 • 1
Visually check upon
receipt

Action if
requirements
are not met
Repair or return
to manufacturer
Repair or return
to manufacturer
Repair or return
to manufacturer
Repair or return
to manufacturer
Determine cor-
rection factor,
or reject
Same as Sec. 3.3
Same as Sec . 3-1
Return to
manufacturer
(Continued)

-------
Table 1.1  (Continued)
                                                                Section No. 3.16.1
                                                                Date June 30, 1988
                                                                Page 12
Apparatus
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Evacuated Con-
tainer and Adsorp-
tion Tube
Sampling

Tedlar bag
Constructed of
material in which
organics remain
stable and are not
retained; leak free
Upon receipt, leak
check and conduct
stability and
retention check
 Return to man-
 facturer, change
 material, or
 use different
 sampling tech-
 nique
Rigid leak-
proof container
Leak free and
of proper size
Upon receipt,
visually check and
then conduct leak
check
 Repair or return
 manufacturer
Pump
Leak free and of
proper capacity
Visually check and
then conduct leak
check and flow rate
check
 Repair or return
 to manufacturer
Flowmeter
Proper flow rate
range and cali-
brated
Upon receipt, check
specifications, check
visually, then
calibrate
 Return to manu-
 facturer or
 repair and then
 recalibrate
Adsorption
tube
Proper material,
adequate capacity,
and consisting of a
primary and second-
ary section
Conduct laboratory
evaluation or consult
literature
 Replace or make
 modification and
 recheck
Personnel sampling
pump
Proper flow rate
range and calibrated
Upon receipt, check
specifications,  then
calibrate
 Return to manu-
 facturer or
 repair and then
 recalibrate
(Continued)

-------
Table 1.1  (Continued)
                                                                Section No. 3.16.1
                                                                Date June 30, 1988
                                                                Page 13
Apparatus
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Direct Pump
Sampling

Pump
Stainless steel or
Teflon-lined, proper
capacity, leak free,
and heated
Visually check, then
conduct leak check,
flow rate check, and
system heating check
 Return to manu-
 facturer or
 repair and
 recalibrate
Flowmeter
Proper flow rate
range, leak free,
and heated
Visually check, then
conduct leak check,
flow rate check, and
heating check
 Return to manu-
 facturer or
 repair and
 recalibrate
Explosion Risk
Bag Sampling

Nonexplosive
vacuum source
Proper flow rate
capacity and
intrinsically safe
Check with plant
safety rules and
check flow rate
capacity
 Return to manu-
 facturer or
 repair and
 recheck
Heated Bag
Sampling

Sampling bag
 Same as above
 Same as above
  Same as above
Heated bag
container
 Leak free, adequate
 capacity, and
 heat system
 capable of main-
 taining proper
 temperature
 Visually check,
 then conduct
 leak check and
 heating check
  Return to manu-
  facturer or
  repair and
  recheck
Heated sample
lines and probe
 Constructed of
 Teflon and/or
 stainless steel
  Visually check,
  then conduct
  heating check
  Return to manu-
  facturer or
  repair and
  recheck
Prefilled
Bag Sampling

Heated flowmeter'
(Continued)
 Same as above
  Same as above
  Same as above

-------
                                                                Section No. 3-16.1
                                                                Date June 30, 19*88
                                                                Page 14
Table 1.1  (Continued)
Apparatus
Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Stainless steel or
Teflon-lined pump
Same as above
  Same as above
  Same as above
Heated box for
flowmeter and
pump
Proper flow rate
range and capacity;
heating system
capable of main-
taining the proper
temperature
  Visually check,
  then conduct
  leak check, flow
  rate check, and
  heating check
 Return to manu-
 facturer or
 repair and
 recheck
Direct
Interface
Sampling
Heated probe, pump
and sample
lines
Same as above
Same as above
Same as above
Heated GC sample
valve
Proper valve and
heating system;
consult owner's
manual
Visually check,
then conduct
check of heating
Return to manu-
facturer or
repair and
recheck
Dilution
Interface
Sampling

Stainless steel or
Teflon-lined pump
Same as above
  Same as above
  Same as above
Dilution pump
Teflon-lined
metering pump
with capacity
of 150 cc/min
Visually check,
then calibrate
 Return to manu-
 facturer or
 repair and
 recalibrate
Flowmeters
Proper flow rate
range and
calibrated
Visually check,
then calibrate
 Return to manu-
 facturer or
 repair and
 recalibrate
Diluent gas
Hyrocarbon-free
air, nitrogen, or
dry cleaned air
Visually check cylin-
der; check cylinder
pressure; run a blank
to monitor impurities
 Return to
 manufacturer
(Continued)

-------
                                                                Section No. 3.16.1
                                                                Date June 30, 1988
                                                                Page 15
Table 1.1  (Continued)
Apparatus
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Heated box for
sample dilution
system
 Heating system with
 temperature con-
 troller and monitor
Conduct heating
check
Return to manu-
facturer or
repair and
recheck
Sample Analysis

Gas chromatograph
 Suitable detector,
 precision of ^ 5%,
 and accuracy of
 + 10%
Refer to Table C
in Method Highlights
then check GC with
applicable organics
Return to manu-
facturer or
repair and
recheck
GC column
Adequate peak
resolution
 Upon receipt, use
 procedure described
 in 40 CFR 60, App. C
 or Method 625
Return to manu-
facturer or
change con-
ditions and
recheck
Strip chart re-
corder or elec-
tronic integrator
See owner's manual
 Upon receipt, check
 as recommended by
 manufacturer
Repair or
return to
manufacturer
Regulators
Proper CGA fittings
and pressure
control
 Upon receipt, attach
 to cylinder and leak
 check
Return to manu-
facturer or
repair or
replace fitting
and recheck
Reagents and
Glassware

GC carrier gas
As specified by GC
owners manual and
hydrocarbon-free
 Visually check upon
 receipt; check cyl-
 inder pressure
Return to
manufacturer
Auxiliary gas
As specified by
owners manual
 Visually check upon
 receipt; check cyl-
 inder pressure
Return to
manufacturer
Calibration gas
(Continued)
Proper compounds
and known concen-
tration in proper
range
 Upon receipt, check
 cylinder tag and
 certification; check
 cylinder pressure
Return to manu-
facturer or
remake or rename

-------
                                                                Section No. 3-16.1
                                                                Date June 30, 19£8
                                                                Page 16
Table 1.1  (Continued)

Apparatus
Zero gas

Acceptance limits
Hydrocarbon- free
air or nitrogen

Frequency and method
of measurement
Visually check upon
receipt; check cyl-
inder pressure

Action if
requirements
are not met
Return to
manufacturer

-------
                                                              Section  No.  3.16.2
                                                              Date  June  30,  1988
                                                              Page   1
 2.0   CALIBRATION OF  APPARATUS

       Calibration of  the  apparatus  is  one  of  the  most  important  functions  in
maintaining data  quality.   The  detailed calibration procedures  included in  this
section were  designed  for  the  sampling equipment specified in  Method  18 and de-
scribed in  the  previous  section.  The  calibration  of the analytical equipment  is
described  in  the  section  detailing  the  analytical  procedures,  Section 3-16.5-
Table 2.1 at the end of this section summarizes  the  quality assurance  functions for
the calibrations addressed in this section.   All calibrations  including  the analyt-
ical equipment should be recorded on standardized forms  and retained in  a calibrat-
ion log book.

2.1  Metering Systems

2.1.1  Net  Test Meter  -  The wet test meter must be calibrated and  have  the  proper
capacity.   For  Method 18,  the wet test  meter  should have a  capacity  of about  1
L/min.  No upper limit is placed on  the capacity; however,  the wet  test meter dial
should make at least one complete revolution at  the  specified  flow rate  for each  of
the three independent calibrations.
       Wet test meters are  calibrated by  the manufacturers  to an  accuracy of +_ 2%.
Calibration of  the  wet test meter must be  checked  upon receipt  and  yearly  there-
after.  A liquid positive displacement  technique  can  be used  to  verify  and adjust,
if necessary, the accuracy  of the wet  test meter to  +_ 2%.  This technique  is de-
scribed in Sections 3-5.2, 3.8.2, and 3-15-2 of  the  Handbook.

2.1.2  Dry Gas  Meter  -  A dry gas meter  is  required  for  gas  and liquid injection
calibrations,  to  prefill bags prior to sampling using the prefilled bag dilution
technique, and/or to  calibrate the flow  rate meters.   For Method  18,  the dry gas
meter is the same size or smaller than  the  dry  gas  meter  typically  used for  Method
6.   The  meter must have  an accuracy  of +_ 3$ for the flow rate  and  sample  volume
used.  Calibration of  the dry gas meter must be  conducted  initially  upon receipt,
quarterly  when  utilized  to make laboratory calibration  standards,  and  following
each field test series for  field use.   The  calibration  procedures are described  in
Section 3-5-2 of this Handbook.

2.1.3  Flow Pate  Meter(s) - Flow rate  meters are needed  for  (1) sampling and (2)
making calibration standards.   Since  they are used  to determine  flow rate and for
total volume determinations, the  flow rate meter(s) selected must have  an accuracy
of  +_ 3%  for  the  flow  rate  and  total sample  volume  for which  they  are   used.
Acceptable flow  rate  meters   include  rotameters,  critical  orifices,   mass
flowmeters, and dry gas  meters.   If  data from  the flow rate meter  is used only  as
an indicator of the flow rate  and is not  used in any  of the emission  calculations,
then the accuracy of +_ 3%  does not  apply.  While it is desirable to  calibrate the
gas flowmeter with the cylinder gas  to be measured,  the  quantity  available and cost
may preclude it.  The error introduced  by using  the diluent gas  for calibration  in
place of the actual gas  to  be measured is insignificant for gas mixtures of  up  to
1,000 to 2,000 ppm.

       Initial Calibration  -  The flow  metering system should be  calibrated  when
first purchased  and at  any  time the  posttest  calibration yields a  calibration
factor  that  does  not agree   within  5%  of  the  pretest   calibration factor.    A

-------
                                                              Section No. 3-16.2
                                                              Date June 30, 1988
                                                              Page  2


calibrated wet  test meter,  calibrated  dry gas meter,  or a properly  sized bubble
meter should be used to calibrate the metering system.
     The flow  rate meter should be  calibrated in the  following  manner before its
initial use in the field.
     1.  Leak check the flow rate meter and pump as follows:
         a.  Temporarily  attach a suitable rotameter (e.g., 0-40 cm3/min)  to
             the outlet of  flow rate system.   The pump  should be placed either
             before or after the flow rate meter based on where  it will be used.
             Place a vacuum gauge at the inlet to the drying tube.
         b.   Plug the flow rate system  inlet.   Evacuate  to a  pressure  at least
             equal to the lowest pressure  that will  be  encountered during use.  If
             the  system  usually operates  at  or near atmospheric conditions,  then
             pull a  vacuum of  25 mm  (1 in.) Hg.
         c.  Note the flow rate as indicated by the rotameter.
         d.  A  leak  of  <0.02 L/min must be recorded  or leaks  >  0.02 L/min must be
             eliminated.
     2.  Attach the  wet  test meter,  bubble  meter,  or calibrated dry  gas  meter to
         the inlet of the flow  rate metering system.
     3.  Run the pump for 15 minutes with the flow rate set at the midrange flow to
         allow the pump to  warm up  and  to permit  the interior surface of the wet
         test meter to become wet.
     4.  Collect  the information  required in the  forms provided  {Figure 2.1A
         (English units) or  2.IB (metric  units) when calibrating a dry gas meter  ,
         rotameter,  or  mass flow meter,  and  Figure  2.2A (English units)  or  2.2B
         (metric units) when calibrating a critical  orifice}  using sample volumes
         equivalent  to  at  least five revolutions of the dry  test meter.   Three
         independent runs must be made.
         a.  For critical orifices,  runs  will  be  conducted  at the single flow rate
             of the orifice  meter.   The  runs  should  be  at three different vacuums
             that are greater than one half an atmosphere (i.e.,  18,  19 and 20 in.
             Hg.).  This is  to  demonstrate  that the  orifice  yields  the same
             flowrate at all critical vacuums.
         b.  For rotameters, mass flow meters,  and dry  gas  meters,  runs will be
             conducted at three  different flow rates  over the  range to be used
             (top, middle, and bottom of range).
      5. Calculate the YA  for each  run  for  the  dry gas  meter,  rotameter  and  mass
         flowmeter or  calculate the  K'  for the  critical orifice as  shown  on the
         data  forms.    Adjust  and  recalibrate  or reject  the dry gas  meter,
         rotameter, or mass flow meter if one or more values of Yt fall outside the
         interval Y _+ 0.03Y, where Y is the average for three runs.   Otherwise, the
         Y (calibration factor)  is  acceptable and is to  be  used  for future checks
         and subsequent test  runs.   The K'  should be within  3%  of  the average for
         all three  runs.   If this  is not true,  reject the orifice or repeat the
         calibration until  acceptable results  are obtained.   The  completed  form
         should be forwarded to the  supervisor for approval,  and then filed in the
         calibration log book.

     Posttest Calibration Check - After  each  field test series,  conduct a calibra-
tion check as described above in Subsection  2.1 concerning  the initial calibration
with the following exceptions:

-------
Date  /
        / // /
       ric p
       _^_^_^  Calibrated by   /4?^       Meter  system  no.	

Barometric pressure, Pm =   SJl^ite	 in.  Hg   Ambient  temperature

                                           ,  dry  gas 	
       Primary  meter  no.
                                     X
                               ,
Type of primary meter:  wet  test

Type of flowmeter calibrated:   rotameter
,  or bubble meter
                                                ,  dry  gas  meter
   ,  or mass flowmeter
Primary meter readings
Initial
reading
(V . ),a
• pi'*
ft3
0
0
0
Final
reading
(V ) a
v vpf ' •
ft3
0.70G,(s
0.1 Obb
1. 0(f04
Initial
temp , ° F
(tpl)
OF
^1
90
•j-o
Final
temp , ° F
(tpf)
op
Co
6.0550
Final
reading
(vs?)b
ft3 or
ft3 /min
O.OIJJ
0.035"3
O.OS30
Initial
temp
(tsi)
oF
(ol
To
^o
Final
temp
(t.f)
op
(,1
70
?0
Press
drop
(Ds),c
in.
H20
0
0
0
Time
min
(9) d
min
40
lJO
2JO
Calibration
factors
(Y,),6
6.W3
o.w.
0.^5-
(Y)
i — .
—
a
-------
                  Calibrated by    A'        Meter system no.   feO ' '(*>   Primary meter no.  £\d IM ~\
Date _
Barometric 'pressure, Pm = 		
Type of primary meter: wet  test 	X,
Type of flowmeter calibrated: rotameter
                              Ktl
                                         mm Hg  Ambient temperature
                                         _, dry gas 	
                                                                    or bubble meter
                                               , dry gas meter
                                                                      or mass flowmeter
Primary meter readings
Initial
reading
(vpi),a
m3
0
0
0
Final
reading
(V ) a
v vpf ' •
*!>•
20-13
to.fr
30- 2(0
Initial
temp , ° F
(tpl)
°C
TJD.S
1O b
•2-0. (*
Final
temp , ° F
(tpf)
°C
ZO.'Z
1.0.5
10 '.(P
Pres
drop
(DP)C
mm
H20
-136
-130
-/-50
Flowmeter readings
Initial
reading

l. rr
                                                                                                                H-
                                                                                                              «-i O
                                                                                                              (D 2!
                                                                                                                O
                                                                                                              UJ •
                                                                                                              O
                                                                                                              -  OJ
                                                                                                              CO
                                                                                                              oo

-------
Date
                  Calibrated by
                                          Meter system no.  \J}~~
Barometric pressure, Pm =    2-1. ^ G-
                                        in. Hg  Ambient  temperature
Primary meter no.
       °F
Type of primary meter: wet test
                                    >C     . dry gas 	
Type of critical orifice: capillary glass     X   ,  needle or  tubing
                                                                , or bubble meter
                                                                              , or adjustable
Primary meter readings
Initial
reading
(vpi),a
ft3
0
D
0
Final
reading
(V).a
ft3
0.7-J3
0.??3
0.??3
Initial
temp,°F

oF
M
70
T-o
Final
temp,°F
(tpf)
op
Of
To
*to
Pres
drop

-------
        ///
Date   ///ft    Calibrated by
Barometric pressure, P_ =   7 (ft
	  Meter system no.   CD ~/Z-    Primary meter no.
mm Hg    Ambient temperature    ,2-Q. <£~     °C
Type of primary meter: wet test
                                    X-      dry gas 	
Type of critical orifice: capillary glass     X  .  needle  or  tubing
                            ,  or bubble meter
                                                                              ,  or adjustable
Primary meter readings
Initial
reading
(vpi),a
L
0
0
0
Final
reading
(V ) a
v P f ' *
L
^^•n^
ZZ-.I3
Z2./2.
Initial
temp , ° F
(tpl)
°C
zo.z
2.0. 1,
2,0. L,
Final
temp,°F
(tpf)
°C
Z.0.^
Z0.lt
2^0. (,
Pres
drop
<°P>
mm
H20
-/3o
-73o
-fbO
Critical orifice readings
Initial
setting
b
L or
L/min
fl-iL-rJ-
fi*ed
f>7<*cL
Final
setting
b
L or
L/min
HIA
s//4
rflA
Press
drop
c
mm
Hg
4-80
570
£50
Time
min
(e),d
min
Zo
££
2J>
Calculated
flow rate
CQ(std)]e
L/min
/. 042,
/.on
/.04Z,
Calibration
factor6
(K'J
J.OZHt,
O.DZ
-------
                                                              Section No. 3-16.2
                                                              Date June 30, 1988
                                                              Page  7


     1.   The  leak check  is  not conducted because  a  leak may have  been corrected
         that was present during testing.
     2. .Three or more revolutions•of the dry gas .meter may be used.
     3.  Only two runs need be conducted at the average flow rate during the test.
     4.  Record  the  calibration check  data on the appropriate posttest calibration
         check  data  form,  Figure  2.2A  (English units)  or  Figure 2.2B  (metric
         units).
     5.  If the posttest Y or K' factor agrees within 5$ of the pretest factor, the
         flow meter .is acceptable.   If the factor does  not agree due  to  a leak,
         correct  the leak  and  recalibrate  the  flow  rate  device.   The  reported
         results should then be calculated using both  the factor obtained with the
         leak and the factor obtained  without  the leak.   If the  flowmeter does not
         pass the  calibration  check,  the  metering system must  be  recalibrated as
         described  above  for  the  initial  calibration.    Either  calculate  the
         emission results  for  the test  report  using  both factors or  consult with
        • the Administrator.

2.1.4   Personnel  Sampling. Pump  -  Personnel .sampling pumps  are used  to collect
samples using  adsorption tubes.   They should  be calibrated before and  after the
field trip using a soap bubble meter as follows:
     1.  Set up the calibration apparatus as shown in Figure 2.3-
     2.  Check  the  pump  battery  with  a voltmeter  to  assure   adequate  voltage;
         charge, if necessary.
     3-  Turn the pump  on and  moisten  the inner surface of the soap  bubble meter
         with  soap  solution;  draw  bubbles  upward until  they  travel  the  entire
         length of the bubble meter without breaking.
     4.  Adjust  the  pump  to desired nominal  flow rate.   Check  the  manometer; the
         pressure drop should not exceed 25mm Hg  (13 in.) water.
     5.  Start a soap bubble and measure the time with a stopwatch that it takes to
         traverse at least 500 ml. Repeat at least twice more.  Average the results
         and calculate  the  flow rate   by  dividing the  calibration  volume  by the
        .average time.
     6.  Record the following data:
         a.  volume measured
         b.  elapsed time
         c.  pressure drop
         d.  air temperature
         e.  atmospheric pressure
         f.  serial number and model of the pump
         g.  date and name of operator
     7-  If the  pump used for  sample  collection uses a  rotameter,  the calibrated
         flow rate must be adjusted for the ambient pressure and temperature during
         sampling:
                         Vc0rr   =  Q 8(Pc T8 / Ps Tc)  °-5
                                                                 Equation 2-17
         where

               V  =  Corrected sample volume, liters,
               Q  =  Indicated flow rate, liters/min,
               6  =  Sampling time,  min,
               Pc =  Pressure during calibration, mm Hg,
               P  =  Pressure during sampling, mm Hg,

-------
                       Tubing
  Soap
 Bubble
  Meter
 (1-Liter)
 Beaker
Containing
  Soap
 Solution
                                                                                         Personnel
                                                                                         Sampling
                                                                                          Pump
                                                                                                                TJ O CO
                                                                                                                to to n>
                                                                                                                m rt- o
                                                                                                                0) (D rt
                                                                                                                  c-i O
                                                                                                                OOC 3
                                                                                                                   O
                                                                                                                   -  l-O
                  Figure  2.3-   Personnel  pump  calibration apparatus.
                                                                                                                   OO
                                                                                                                     ON

-------
                                                              Section No. 3.16.2
                                                              Date June 30, 1988
                                                              Page  9


               T  =  Temperature during calibration, °K, and
               Ts =  Temperature of sample gas, °K.

2.2  Dilution System

2.2.1  Dynamic Dilution System  -  A  dynamic  dilution system may be required for (1)
preparation of low concentration standards from high concentration standards or (2)
for  measuring high  concentrations  of  organic  emissions.    The dynamic  dilution
system must be initially calibrated in  the  laboratory  and then checked during each
use.    To  prepare  the  diluted  calibration  samples,  calibrated  rotameters  are
normally used to meter both  the high  concentration calibration gas  and the diluent
gas.   Other  types of  flowmeters  and commercially available  dilution  systems can
also be used provided they meet the performance criteria described below.
     The following steps  should be used  to conduct the  laboratory  calibration of
the dynamic dilution system:
     1.  Assemble the dilution system (see Figure 2.4)  as a unit using a calibrated
         rotameter or mass flow meter for the  calibration or stack  gas in combina-
         tion with a calibrated rotameter,  mass  flowmeter or dry gas meter for the
         diluent gas.  It is  recommended  for dilutions  up to 20 to  1 that a single
         dilution system be used.  For dilutions greater  than 20  to  1,  a double
         dilution system should be used.  It is also recommended that the system be
         assembled as a unit  and  not  be disassembled between uses.   The rotameters
         should be calibrated for  the  range in  which  they will  be  used following
         the calibration procedures described above.
     2.  Leak  check  the system by plugging  the  inlet  line to both  rotameters,
         placing the dilution system discharge  line in a container  of water, and
         turning on  the sample pump.   The system  is  leakless if no  bubbles are
         released from the discharge line.
     3.  The dilution system can be calibrated over the range  that it will be used,
         however,  if the  exact dilution  to  be used   is known, it  is better to
         conduct a triple calibration at the desired dilution  setting.   Attach the
         dilution system to  the diluent and calibration gases.  Set the flowmeters
         to the desired rate  and  fill the bag  with sufficient gas for GC analysis.
         Be careful not to  overfill the bag and cause  the  bag to apply additional
         pressure on the dilution system.   Record the  flow  rates of both flowmet-
         ers ,  and the  laboratory  temperature  and atmospheric pressure on  the
         dynamic dilution calibration form,  Figure 2.5  or an equivalent form.
     4.  Analyze the diluted  calibration  gas and a  calibration  gas  that is in the
         same range  as  the diluted gas.  The  two gases  must  agree  within 10% for
         the calibration point to be acceptable.   Repeat  the calibration runs until
         acceptable results are obtained at all desired settings.

2.2.2  Static Dilution System - The static  dilution system can be used for (1) the
bag  sampling  technique  and  (2) for preparation of low  concentration  calibration
gases  from  high concentration  cylinder gases.   The dilution method  for  the bag
sampling technique is used  to reduce the concentration of organics  or  water vapor
in a gas sample below the condensation  point or  for safe handling,  below the lower
explosive limit.  Static dilution involves filling a bag with a diluent gas using a
calibrated dry gas meter  or mass flowmeter and  using  a  syringe  or  a rotameter to
add  the  calibration  gas  or a sample  of stack  gases to make  a lower concentration
calibration or sample gas.
     The following steps should be used  to calibrate  a  static  dilution  system in
the laboratory before use:

-------
                                             Vent to Charcoal Adsorbers
Heated Line
from Probe
                Quick
                Connect
                                                              Quick
                                                           Connects to
                                                           Gas Sample
                                                              Valve
                             Source
                            Gas Pump
                            1.5L/Min
                   150 cc/Min
                     Pump
150 cc/Min
  Pump
                                                     3-Way
                                                     Valves
                                                     in 100:1
                                                     Position
Flowmeters
(On Outside
  of Box)

Flow Rate of
1350 cc/Min
 Check Valve
Quick Connects
 for Calibration
                                                                                                                          CO
                                           Heated Box at 120° C or Source Temperature
                                                                                                 To Heated GC Sampling Valve
                          Figure 2.4.   Schematic of  heated box required for  dilution  of samples.
                                                          O CO
                                                          CO (D
                                                          ct O
                                                          CD rt
                                                            H-

                                                          §§
                                                          0> Z
                                                            o
                                                          oo •
                                                          o
                                                          -  00
                                                                                                                            VD ON
                                                                                                                            CO •
                                                                                                                            osro

-------
                                                              Section No. 3.16.2
                                                              Date June 30, 1988
                                                              Page  11
                           P-
                           g-c,
                           AI/A
Date     -*-/ II /0D                     Calibrated by 	
                                        Date source meter calibrated  I/ £&/6G>
                                        Date stage 1 meter calibrated  / / 26>/ 8
                                        Date stage 2 meter calibrated
                                        Heated box temperature
                 	Leak check for total system	
Certified concentration 22,1d>  ppmv(X)  Date of calibration curve    <2-Jl116&
Source flowmeter number
Stage 1 flowmeter number
Stage 2 flowmeter number
Barometric press  /6>/   mm Tin.) Hg
Organic compound
STAGE 1                                          RUN 1
Emission gas flowmeter reading,  ml/min  (qc i}    / 5"d
Diluent gas flowmeter reading, ml/min (qdl)    I OOP
Dilution ratio                                 ~
Injection time, 24h
Distance to peak, cm
Chart speed, cm/min                            AJfA
Retention time, min
Attenuation factor
Peak area or units
Peak area X attenuation factor
Measured concentration,3  ppmv               	
Calculated concentration,15 ppmv  (Cs)           Z6fe
Percent difference,0 %                         -/. 3
                                                              RUN 2
                                                             looo
                                                            1112.
                                                                          JI4A
                                                                          A//A
                                                                           3.5-^
                                                             Z9OO
                                                                          ^600
                                                  2.0V
                                                            II . (sDO
                                                  •&
                                                                         ZSft
                                                                        -*- 0. 3
STAGE 2 (if applicable)
Emission gas flowmeter reading, ml/min  (qc2)
Diluent gas flowmeter reading, ml/min (qd2)
Dilution ratio
Injection time, 24h
Distance to peak, cm
Chart speed, cm/min
Retention time, min
Attenuation factor
Peak area or units
Peak area X attenuation factor
Measured concentration,3 ppmv
Calculated concentration,d ppmv
Percent difference,0 %
RUN 1
Ail A-
RUN2
AllA
RUN3
A//A
                                                 \
  See Figure 5-1 for calculation.

       106 x (X x qc)
                      = Calculated concentration for single stage

                       Calculated Concentration - Measured Concentration
0 Percent Difference =
d C  = 106 x X
                                                                         x 100%
                                    Measured Concentration
                               x
                                             ,)
                                                 = Calculated cone, for two stage
                                   ^02 + 3d

                      Figure 2.5- Dynamic dilution data form.

-------
                                                              Section No. 3.16.2.
                                                              Date June 30, 1988
                                                              Page  12


     1.  Assemble  the static dilution  system  (see Figure 4.3)  and  leak check the
         system  by plugging  inlet  to  the  dilution system, placing  the discharge
         line in a container of water, and pulling  a  vacuum of about 1 in. of Hg.
         The system is leakless if no bubbles are  released from  the discharge line.
         If the  system is not leakless, find the leak  and correct it.
     2.  Calculate  as  the  amount  of  diluent  gas  needed  to  obtain  the  desired
         dilution  or  calibration gas  concentration.   Meter the desired  amount of
         gas into  the bag.   If the purpose of  the static dilution  is to prepare a
         lower  concentration of  calibration gas,  the  calibration  gas  should be
         added to the bag using a gas tight syringe.   Record the data on the .static
         dilution  system  data form,  Figure 2.6 or similar  form.  If  the purpose of
         the  static  dilution  system  is  to collect  a  diluted stack  sample,  the
         calculated amount of diluent  gas  is added to  the bag and the stack gas is
         metered  into the  bag  from  the  stack.    To calibrate this  system,  the
         calculated amount of diluent gas should be metered into the bag and then a
         calibration  gas  should be metered into the bag  with the flowmeter that is
         to be  used in the  field.   Record the data on  the  static  dilution system
         data form, Figure 2.6 or similar data  form.
     3.  Analyze  the  diluted calibration  gas  and analyze  a  different calibration
         gas that  is  in the  range of  the  diluted calibration  gas.   The two gases
         must agree within 10% of each other for the system to be acceptable.

2.3  Thermometer

     The thermometers(s)  on  the metering  systems  and the sample probes  and lines
should be initially compared with a mercury-in-glass thermometer that meets ASTM E-
1 No. 63C or 63F specifications:
     1.  Place  the thermometer  to  be calibrated  and  the  mercury-in-glass thermo-
         meter  in  a bath of  boiling water.   Compare  the readings  after  the bath
         stabilizes and  then record  on the calibration  data  form.   Figure  2.7 or
         equivalent.
     2.  Allow both thermometers to come to room temperature.  Compare the readings
         after the thermometers stabilize.
     3-  The  thermometer is  acceptable if the values agree within  3°C (5-4°F) at
         both points.
     4.  Prior  to  each  field  trip,  compare the temperature  reading of the mercury-
         in-glass thermometer at room temperature with that of the thermometer that
         is part of the metering system.  If the values are not within 6°C (10.8°F)
         of each other, replace or recalibrate  the meter thermometer.

2.4  Barometer

     The field  barometer  should be adjusted initially and  before each test series
to agree within 2.54 mm (-0.1 in.) Hg with a mercury-in-glass barometer or with the
pressure value reported from a nearby National Weather Service Station and correct-
ed for  elevation.   The  tester  should be  aware that  the National Weather Service
readings are  normally  corrected  to  sea  level;  uncorrected  readings should  be
obtained.  The  correction for the elevation  difference between the  weather station
and  the  sampling point should be applied  at a rate of  -2.5  mm Hg/30  m  (-0.1  in.
Hg/100 ft)  elevation increase, or vice versa for elevation decrease.

-------
                                                           Section No.  3.16.2
                                                           Date June  30,  1988
                                                           Page   13
Date        ,   .	
Source flowmeter number    p-
Dry gas meter number
Ambient temperature
Barometric press
Organic compound
Certified concen. (X)  2-2-/Q  I
                     	  'C  (°F)
                     mm (in.) Hg
                            ppmv
                                        Calibrated by 	
                                        Date source meter  calibrated  _
                                        Date dry gas meter calibrated
                                        Dry gas meter calib factor  (Y)
                                        Leak check for  total system 	
                                        Vacuum during leak check  	^
                                        Date of calibration curve    p-
                                                                            83
Initial 'dry gas meter reading, L  (ft3)
Final dry gas meter reading, L (ft3)
Volume of diluent gas metered, L  (ft3)
Gas metered X calibration factor  (Y),{V2
Flowmeter sampling rate, L/min (cfm)
Sampling time, min
Sampling rate X sample time, L (ft3),{V1
Dilution ratio
Injection time, 24h
Distance to peak, cm
Chart speed, cm/min
Retention time, min
Attenuation factor
Peak area or units
Peak area X attenuation factor
Measured concentration,3 ppmv
Calculated concentration,1" ppmv,  {Cs}
Percent difference,0 %
                                             RUN 1
                                                             RUN  2
                                           /3 Z. 31 (,
                                                            /3>-*>
                                             O. 151-
                                              2.O
                                                                         z-o
                                            7.4-12
                                                            8>.//0
                                                           2-6, (*
                                               D
                                                          10
                                           I J- 2-00
                                                             //'Z-66
                                                          3-3-0
                                                          - /.&
a  See Figure 5-1 for calculation.

b  Calculated concentration  (Cs) =
                                      X  (V,)
                                           V)
                                                                    ppmv
Percent difference, %d =
                             Measured concent - Calculated concent

                                     Measured concentration
                                                                     X  100  =
The percent difference must be less than 10 % absolute.
                   Figure 2.6.  Static dilution data form.

-------
Date
bv/ffi
/iv/M






Reference
thermometer
type
A S7"A^
6$F
forty
60 F






Calibr
thermc
type
£3T
CeufiUi,
f





ated
meter
use
5^
hj&kr-
bo?






no.
TAJ -12.
9-1






Ambier
refer8
rfr
-dl'F






it temper
calibrb
7l°f
lor






Measured
•ature
differ0
*2V
*-!'/=






values
Bod
refer8
1I1°F
urF






ling wat
calibr5
*a-F
3.11'F






er
differc
W
-/ •/=






Calibrator's
initials
ATM
A771/






•now
(B 03 (D
oq ct n
(D (0 ct
    p-
  t-i O
*-* c y
4=- 3
  ID Z
    O
  VD O\
  00 •
  COM
Temperature reading of the reference  thermometer in °C or °F.

Temperature reading of the thermometer being calibrated in °C or °F.

Difference between the reference  thermometer and the calibrated thermometer.  This difference  must

be less than 3°C  (5.4°F)  for  than initial calibration and 6°C (10.4°F) for the calibration check.


                            Figure 2.7-  Thermometer calibration form.

-------
                                                              Section No.  3.16.2
                                                              Date June 30, 1988
                                                              Page  15
             Table  2.1.   ACTIVITY  MATRIX  FOR  CALIBRATION OF EQUIPMENT
Apparatus
 Acceptance limits
Frequency and method
   of measurement
                                              Action if
                                              requirements
                                              are not met
Wet test meter
Capacity of about 2
L/min and accuracy
within 1%
 Calibrate initially,
 then yearly by
 liquid displacement
                                              Adjust until
                                              specs are met, or
                                              return to vendor
Dry gas meter,
mass flow meter,
and rotameters
Y.  = Y + 0.03Y at a
point greater than
the flow rate range
to be used
 Calibrate vs. wet,
 dry, or bubble meter
 upon receipt and
 after each test
                                              Repair and
                                              then recalibrate,
                                              or replace
Critical orifice
K'i = K ± 0.03K'
                       Calibrate vs.  wet,
                       dry, or bubble meter
                       upon receipt and
                       after each test
                        Repair and
                        then recalibrate,
                        or replace
Dilution system
Measured value for
diluted and undi-
luted calibration
gas must agree
within 10%
 Calibrate upon
 receipt and prior to
 each field test using
 calibration gases
                                              Correct problem
                                              and rerun cali-
                                              bration
Thermometers
Within 3°C (5.
of true value
 Calibrate initially
 as a separate com-
 ponent with mercury-
 in-glass thermometer;
 check before each
 test against mercury-
 in-glass thermometer
                                              Adjust or replace
Barometer
Within 2.5 mm
(0.1 in.) Hg of
mercury-in-glass
barometer or weather
station value
 Calibrate initially
 using mercury-in-
 glass barometer;
 check before and
 after each test
                                              Adjust to
                                              agree with
                                              certified
                                              barometer

-------
                                                                 Section  No.  3-16.3
                                                                 Date  June  30,  1988
                                                                 Page  1


3.0   PRESAMPLING OPERATIONS

       This section addresses two major  areas of presampling operations  for  Method
18:  (1)  preparation  for,  performance of,  and  interpretation of  results  for  the
preliminary survey  and  (2)  preparation  for the final  sampling.   This section  de-
scribes the preliminary survey  as  it applies  to Method 18; for  additional general
information on preliminary surveys, see  Section 3-0 of this Handbook.  The quality
assurance  activities  for  the  preliminary  survey  activities  and  the presampling
activities for final testing are summarized in Tables  3-1  and  3-2,  respectively,  at
the end of this section.

3-1  Preliminary Survey Measurements

       The preliminary survey measurements are needed to  properly  design the final
emission test sampling and analysis protocol.  The  primary objective of the prelim-
inary survey is to collect a preliminary survey  sample for determining which  sampl-
ing procedure  is  most appropriate and  for developing the optimum analytical  pro-
cedures.   Using  the preliminary survey sample,  estimates of the source  concentra-
tion are made  and the major organic  components in the gas stream are identified.
Also,  any  compounds that  may interfere  with  the quantitation of the target anal-
yte(s) are identified and the appropriate  changes  in  the  analytical procedures  are
made.  Other measurements made  during the  preliminary survey  include  sampling site
dimensions and gas  stream  properties.   The preliminary  survey is  also  used  to
obtain a description of the process being sampled,  to  determine sampling  logistics,
and,  when  possible,  to  collect bulk process samples and  use emission screening
techniques.  Use the data form shown in Figure 3-1  to  record the preliminary  survey
information.

3.2  Preliminary Survey Preparation

     This  section addresses  the equipment and  preparatory  activities  needed  to
conduct the preliminary  survey.  Figure 3-2  can serve as an  equipment  checklist,
packing list,  and/or equipment status form for the  preliminary survey.

3.2.1  Measurement  of Flue  Gas  Properties - The apparatus that may be required to
supplement information obtained  from  plant personnel during the  preliminary  survey
concerning  the moisture  level, temperature,  and  static  pressure of  the  source
should be prepared for the preliminary survey as  follows:

     Barometer -  The field  barometer should be compared with a  mercury-in-glass
barometer  or with a National Weather Service Station (see Subsection 2.4) reading
prior to each field test.

     Met Bulb/Dry Bulb Thermometers - It is recommended that for sources  with stack
temperatures at or  below 59° C,  wet bulb/dry bulb thermometers  be used to determine
stack gas  moisture  content. The  thermometers  should be compared with  a mercury-in-
glass  thermometer at  room  temperature prior  to  each  field trip.  The wet  bulb/dry
bulb measurement may also  be used,  with the prior approval of  the Administrator,  to
determine stack gas moisture for sources where the  stack  temperature exceeds  59°C.

     Method 4 Equipment - For sources with stack temperatures  above 59°C, Method k
equipment is  recommended  to determine stack gas moisture content.   Prepare  the

-------
                                                                 Section No. 3.16L.3
                                                                 Date June 30, 1988
                                                                 Page 2
        I.  Name of company 	 Date_
            Address
            Contacts                                 Phone
            Process to be sampled
            Duct or vent to be sampled_
        II. Process description
            Raw material
            Products
            Operating cycle
                 Check:  Batch	Continuous	Cyclic
                 Timing of batch or cycle	
                 Best time to test	
        III.  Sampling site
             A.  Description
                 Site description	
                 Duct shape and size_
                 Material
                 Wall thickness	inches
                 Upstream distance	inches	diameter
                 Downstream distance	inches	diameter
                 Size of port	
                 Size of access area	
                 Hazards	Ambient temp	°F
                 Properties of gas stream
                 Temperature        °C	°F,   Data source	
                 Velocity	,  Data source	
                 Static pressure	inches H20,  Data source_	
                 Moisture content	%,  Data source_	
                 Particulate content	,  Data source	
                 Gaseous components
                    N2	%  Hydrocarbons (ppm)    Toxics/Acids (ppm)
                    02	%  	     H2S 	
                    CO	%  	     HC1 	
                    C02	%  	     HF  	
                    SO.,           %                        Other
                    Figure 3-1.  Preliminary survey data sheet.

(Continued)

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 3
Figure 3-1 (Continued)

                 Hydrocarbon components
                                                            ppm
                                                            ppm
                                                            ppm
                                                            ppm
                                                            ppm
                                                            ppm
             C.  Sampling considerations
                 Location to set up GC	
                 Special hazards to be considered_
                 Power available at duct	
                 Power available for GC	
                 Plant safety requirements
                 Vehicle traffic rules
                 Plant entry requirements_
                 Security agreements_
                 Potential problems
                 Safety equipment (glasses, hard hats, shoes, etc.)
             D.  Site diagrams.  (Attach additional sheets if required).

        IV. On-site collection of preliminary survey samples
             A.  Evacuated flasks
                 Flasks have been cleaned, heated in furnace and purged
                 with nitrogen?	
                 Flask evacuated to the capacity of pump?	
                 Filter end of. probe placed at center of stack, probe
                 purged and sampled collected into flask until flask is at
                 stack pressure?	
                 Stopcocks closed and taped?	
                 Duct temperature and pressure recorded?	
                 Purged flasks
                 Flasks cleaned and purged with nitrogen?	
                 Filter end of probe placed into stack, sample purged for
                 2 to 5 min and then stopcocks closed?	
                 Stopcocks taped to prevent leakage?	
                 Duct temperature and pressure  recorded?
                 Stability and adsorption checks conducted?

(Continued)

-------
                                                                 Section No. 3-16..3
                                                                 Date June 30, 1988
                                                                 Page 4
Figure 3-1  (Continued)
             C.  Flexible bags
                 Bags have been blanked checked and leak checked?
                 Sampling system leak checked?	
                 Filter end of probe placed into center of stack and sample
                 obtained at a proportional rate for appropriate amount of
                 time?	 •
                 Duct temperature, barometric pressure, ambient temperature,
                 flow rate, static pressure, and initial and final sampling
                 time recorded?	
                 Analysis performed within 2 hr?	
                 Stability and adsorption checks conducted?	
             D.   Adsorption tubes
                 Proper adsorption tube(s)  selected based on the likely
                 analytes?	
                 Probe or adsorption tube placed into center of stack and
                 sample obtained at a constant rate with a calibrated
                 system for appropriate time based on the expected concen-
                 trations of analytes?	
                 Total sample time and sample flow rate (or the number of
                 pump strokes),  the barometric pressure,  and ambient
                 temperature recorded?	
                 Water vapor was less than 2% or measures were taken to
                 protect or increase the adsorption capacity of the
                 adsorption tube(s)?	
             E.   Quality assurance performance audit samples
                 Quality assurance audit samples collected in the same
                 manner as the emission samples?	
             F.   Bulk samples and screening techniques
                 Bulk emission sample(s)  collected?	
                 Bulk liquid sample(s)  collected?	
                 Detector tubes or other screening techniques used?

-------
                                                                 Section No.  3.16.3
                                                                 Date June 30,  1988
                                                                 Page 5
Apparatus check
Moisture Determination
W Bulb/D Bulb
Checked
Barometer
Calibrated*
Method 4
Probe, heated &
leak checked
Impingers
Meter system
calibrated*
Velocity Determination
Pi tot Tube
Number
Length
Pressure Gauge
Manometer
Other
Stack Thermometer
Calibrated

Evacuated Flask
Evacuated Flasks
Number
Cleaned
Oven heated
N2 purged
Probes
Number
Cleaned
Glass wool
Suction bulb
Pump
Purged Flask
Flask
Number
Cleaned
Oven heated
N2 purged
Acceptable
Yes




No




Quantity
Required




Ready
Yes




No




Loaded and Packed
Yes




No




*Most significant items/parameters to be checked.

                   Figure 3-2.  Preliminary survey preparations
(Continued)

-------
                                                                  Section No.  3-16.3
                                                                  Date June 30,  19*88
                                                                  Page 6
Figure 3-2  (Continued)
Apparatus check
Purged Flask (continued)
Probe
Number
Cleaned
Glass wool
Vacuum Source
Pump
Squeeze bulb

Bag Sampling
Probe Liner
S steel
Glass
Teflon tube
Length
Meter System
Flowmeter*
Pump
Evacuated can
Charcoal tube
Sample line
Tedlar Bags
Number
Blank checked
Leak checked*
Heated Box
Number
Heat checked
Adsorption Tube
Probe
Heated
Checked
Nonheated
Glass
S steel
Filter
Sample Line
Tupe
Length
Checked*
Acceptable
Yes



No



Quantity
Required



Ready
Yes



No



Loaded and Packed
Yes



No



*Most significant items/parameters to be checked.

(Continued)

-------
                                                                 Section No.  3-16.3
                                                                 Date June 30,  1988
                                                                 Page 7
Figure 3-2  (Continued)
Apparatus check
Adsorption Tube (continued)
Pump and Meters
Pump
Orifice
Calibrated*
Rotameter
Calibrated*
Timer
Adsorption Tubes
Type

Bulk Samples
20-ml Jars
Cleaned
Acceptable
Yes


No


Quantity
Required


Ready
Yes


No


Loaded and Packed
Yes


No


*Most significant items/parameters to be checked.

-------
                                                                 Section No. 3-16.3
                                                                 Date June 30, 1<588
                                                                 Page 8

equipment for sampling  following the procedures described in Section 3-3-3 of this
Handbook.  Method 4  equipment may also be used to determine the stack gas moisture
for sources where the stack temperature is at or below 59°C.

     S-Type Pitot Tube  and Differential Pressure Gauge  - Prepare the S-type pitot
tube and the differential  pressure gauge  for sampling following the procedures de-
scribed in Section 3-1-3 of this Handbook.

3.2.2  Glass Flask  Sampling - The apparatus and  reagents required for the collec-
tion  of  preliminary  survey samples  using glass sampling  flasks are  prepared as
described below-  Alternative equipment found suitable  may  be  used  subject to the
approval of the Administrator.

     Probe - If a heated probe is to be used for sampling, then the probe's heating
system should be checked to see that it is operating properly.  The probe should be
cleaned internally by brushing first with tap  water,  then with deionized distilled
water, and finally with acetone.   Allow the  probe to air dry.  The probe should be
sealed at  the  inlet end and  checked for  leaks by applying  a vacuum  of 380 mm (15
in.) Hg.   See Subsection  1.0 for leak check  procedure.  The  probe  is considered
leak  free  under these conditions  if no loss  of  vacuum is  seen  after one minute.
Any  leaks  detected  should be corrected  or  the probe  should be  rejected.  If the
probe has an external sheath,  the integrity  of the  seal between the sheath and the
probe liner should be checked  to ensure ambient air does not dilute the gas sample.

     Teflon Tubing  - Prepare sections of  tubing  for  connections  between the probe
and each flask  (or  bag or tube)  that  constitutes a preliminary survey sample col-
lection  device.   Clean the  tubing using the procedure  described  above  for the
probe.

     Quick Connects  - The  quick  connects  should be  new or cleaned according to the
manufacturer's  recommendations.    Leak check  the quick  connects as  described in
Subsection 1.0.

     Glass  Sampling Flasks  -  Prepare the  glass sampling  flasks for collecting
preliminary survey samples as  follows:  Remove the  stopcocks from both ends of the
flasks, and wipe the parts to  remove any grease.  Clean the stopcocks, barrels, and
receivers with chloroform.  Clean all  glass  parts with a soap solution, then rinse
with tap water  followed  by deionized distilled water.   Place  the flasks in a cool
glass annealing furnace and heat  the furnace to 550°C.   Maintain the flasks in the
oven  at  this  temperature  for one hour.   After  one  hour,  shut  off and  open the
furnace  to  allow the flasks  to cool.   Return the Teflon stopcocks  to  the glass
flasks (if glass stopcocks are used,  apply a light  coating of vacuum grease to the
stopcocks before  returning to the flasks.)   With both  stopcocks  open,  purge each
assembled flask with high purity nitrogen  for 2 to 5 minutes.  Close off the outlet
stopcock followed  by the  inlet  stopcock  to maintain  a slight  positive  nitrogen
pressure in the flask.  Secure the stopcocks with tape to prevent them from opening
accidentally.

     High-Vacuum Pump - A  high-vacuum  pump will be  required for preliminary survey
sample collection using the evacuated  flask  procedure.   Check the operation of the
pump prior to going  to  the field as follows:  Check  for minimum  pump  vacuum of 75
mm (3 in.)  Hg absolute by attaching a Hg-filled U-tube manometer to the pump inlet

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 9

and turning on  the  pump.   If the minimum vacuum cannot be  reached,  then repair or
replace the pump.

     Rubber Suction  Bulb  -  A rubber suction bulb will be  required for preliminary
survey sample collection using the purged flask procedure.   The rubber suction bulb
should be  checked for  proper operation prior  to going to the field  as follows:
Attach the bulb  to  a water manometer, or equivalent, and  squeeze  the bulb until a
vacuum of at  least  250  mm (10 in.) H20 is reached.   Repair or replace the squeeze
bulb if sufficient vacuum cannot be developed.

3.2.3  Flexible Bag Procedure - A flexible bag of Tedlar or aluminized Mylar can be
used to collect preliminary  survey  samples.   If it  is anticipated that Tedlar bags
will be  selected as  the  final  sampling  method,  then it  is  recommended that the
preliminary survey  sample be  collected using  a  Tedlar  bag.   In addition  to the
apparatus described in Subsection 3-2.2 for the glass flasks (with the exception of
the flasks  and the high-vacuum  pump)  the  apparatus listed below  will be required
and should be prepared as follows:

     Tedlar or Aluminized Mylar Bags  -  Prepare new  bags for  preliminary survey
sampling by leak checking the bags before going to the field.  The bags should also
be  checked  for  contamination by filling  with hydrocarbon-free  air  or nitrogen
during the leak  check.   The bags are checked as  follows:   Connect a water manome-
ter, or equivalent, using a tee connector,  between the check valve quick  connect on
the bag and a pressure  source (or hydrocarbon-free  air or  nitrogen  for  conducting
the contamination check).   Pressurize the bag to 5 to 10 cm (2 to 4 in.)  H20.  Loss
of pressure over a 30-second period indicates a leak.  Alternatively, leave the bag
pressurized overnight:  a  deflated bag  the  following day  is indicative  of a leak.
Reject or repair  any bags with leaks.  After the hydrocarbon-free air or nitrogen
has remained  in  the bag for 24  hours,  analyze  the  bag contents using a GC with a
flame ionization  detector on the most sensitive setting.   The bag  should be re-
jected  if  any  organic  compounds  are  detected.    If  any  organic   compounds  are
detected,  the bags may be  used  if  they  are not the  compounds to be  sampled and
analyzed.

     Etgid Leak-Proof Containers -  Rigid containers can be used with the bags for
collecting preliminary  survey samples.  The  rigid containers  used to  contain the
Tedlar bags  during  sampling should  be  checked  for leaks prior  to going  to the
field.   The container  should  be leak  checked  with the bag in place  as follows:
Using a tee connector,  connect a water  manometer  or  equivalent,  between a pressure
source and the  container  outlet.  Pressurize the container to 5 to  10  cm (2 to 4
in.) Hg.   Any loss of pressure after 30 seconds indicates  a leak.  Reject or repair
the rigid container if a leak is indicated.

     Direct Pump  Sampling System -  A direct pump sampling system can  be used in
place of the  rigid containers for  collecting preliminary survey  samples.  If this
method is selected,  then  the system should be  assembled and leak  checked prior to
going to the  field  as follows:   Assemble the system  (see Figure 4.5).   Bypass the
Tedlar bag and its protective container by attaching the  vacuum line directly after
the rotameter  using the quick connects  on the sample and vacuum lines.  Plug the
probe  inlet   and  turn  on the  vacuum  pump.    If  the  system is leak  free,  the
rotameter should eventually  indicate  no  flow.   Alternatively,  the  sample line that
is attached to the sample bag can be  placed  in  water.  If  bubbling stops, then the
system is  leak free.

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 19B8
                                                                 Page 10

     Needle Valve  and  Rotameter - Prior to  each  field trip or at  any  sign of er-
ratic behavior,  the  flow control valve and  the rotameter  should  be cleaned accor-
ding to the maintenance procedure recommended by the manufacturer.

3.2.4  Adsorption  Tube Sampling - The adsorption  tube sampling procedure can also
be used  to  collect the preliminary  survey  sample.  If it  is  anticipated that ad-
sorption  tubes  will  be  selected as  the final  sampling method, then it  is recom-
mended that the preliminary survey samples be collected using tubes containing each
potential type  of adsorbent.  In addition  to the apparatus described  in Section
3.2.1 for the glass  flasks (with the  exception  of  the  flasks and  a  high-vacuum
pump) the apparatus  listed below will be  required and should,  be  prepared as fol-
lows :

     Adsorption Tubes - Check to see that the proper type of tube has been obtained
for collecting the target organic compounds.  Refer to Table B in the Methods High-
lights Section to  determine  the proper adsorption material.  Check to see that the
supply of adsorption tubes is  sufficient  to conduct  the  emission test,  including
field blanks and desorption efficiency determinations.           .  ..     •

     Personnel Sampling Pump -  A personnel sampling  pump  is used to  collect the
adsorption  tube  samples.   The  pump  should be calibrated  following the procedures
described in Subsection 2.1.4.

     Extraction Solvents  - An  extraction solvent will be  required to  prepare the
preliminary survey adsorption tube sample(s) for analysis.  Refer to Table B in the
Methods Highlights Section to determine the proper extraction solvent.

3-3  Preliminary Survey Sample Collection

     The  preliminary  survey sample  collection  includes flue gas  or  duct moisture
and velocity  determinations in  addition to  collection of  actual  flue  gas or duct
samples.                                      .                           .

3-3-1  Preliminary Survey Moisture  Determination  - If  the  moisture content of the
flue gas  in the  duct to be tested cannot be obtained  from  the plant  personnel,  it
is determined  using either  wet  bulb/dry  bulb thermometers  or Method 4 sampling
apparatus,  depending  on  the  flue gas temperature.    If  the flue  gas  temperature
cannot be obtained from plant  personnel,  then determine the  flue gas  temperature
using a  calibrated thermocouple, thermometer,  or  equivalent  temperature  measuring
device.

     Net  Bulb/Dry  Bulb  Procedure -  For  flue gas' streams at  or  below  59°C,  the
moisture  content  of the flue  gas  should be  determined using wet  bulb/dry  bulb
thermometers  and  the  partial  pressure equation shown  below.    Obtain the  wet
bulb/dry bulb temperatures as follows:
     1.  Moisten the wet bulb thermometer wick with deionized distilled water.
     2.  Insert the thermometers into the  flue gas stream and  monitor the wet bulb
        temperature.
     3.  When the wet bulb  temperature has stabilized,  record both  the wet bulb and
        dry bulb thermometer temperatures.
     4.  Calculate the flue gas moisture content using the equations below.

-------
                                                                 Section No. 3.16-3
                                                                 Date June 30, 1988
                                                                 Page 11

               10(6.69H-(3iA4/(Tw + 390.86)»                           Equation 3-1
        w2 =
                    Pb
        %U20 = w2 - (0.00036? x (Td-Tw) x (l+(Tw-32)/157D) * 100      Equation 3-2

     where

        w2 = Calculated constant,  saturation % H20 at Tw ,
        Tw = Wet bulb temperature, °F,
        Td = Dry bulb temperature, °F,
        Pb = Barometric pressure,  in. Hg, and
        Ps = Static pressure of duct, in. H20.

     Method 4 Moisture  Procedure  -  Follow the procedure for Method  4 described in
Section 3-3 of this Handbook.

     Method 2 Velocity  Procedure  -  Follow the procedure for Method  2 described in
Section 3-1  of  this  Handbook  to  determine  the  flue gas  or  duct velocity  at the
sampling  point.    If  the velocity  varies by  more than  10$  during  the projected
sample run  time,  then proportional  sampling will be required as  described in Sub-
section 4.0.  Because of the small  size of  some ducts, Methods  2A,  2C,  or 2D may
have to be  used.   Follow the  criteria and procedures  described  in  the applicable
method.

3-3-2   Collection of Samples with  Glass Sampling Flasks -  Using  the precleaned
glass sampling flasks, preliminary survey samples are collected using the evacuated
flask procedure or the purged flask procedure.

     Evacuated  Flask Procedure   -  Collect  preliminary survey  samples  using the
evacuated flask procedure as follows :
     1.   Using  a high-vacuum pump  which is  connected to one stopcock  while the
          other  stopcock remains closed, evacuate  each  precleaned flask  to the
          capacity of the pump.   A mercury manometer can  be  connected between the
          pump  and the flask  using a tee connector  to indicate  when  the maximum
          vacuum  is achieved.  At this point,  record  the  vacuum,  and close off the
          stopcock leading to the pump.
     2.   Remove  the  tubing  leading to the pump and attach a  glass  tee  (6-mm out-
          side  diameter,  or  equivalent)   to  the  flask  inlet with  a  short piece of
          Teflon  tubing.
     3.   Connect  the end of  the sampling  probe  to the  glass  tee  using a short
          length  of  Teflon  tubing.    The tubing must  be  of sufficient  length to
          reach the sampling point  at the centroid of or  no closer than 1 meter to
          the duct wall.
     4.   Connect  the rubber suction  bulb to  the third  leg of the tee with a piece
          of Teflon tubing or suitable flexible tubing.
     5-   Place a plug  of glass wool in  the  probe inlet,  enlarged to approximately
          12-mm outside  diameter, to serve as a  filter to remove particulate mat-
          ter.
     6.   Place  the  inlet (filtered)  end of  the  probe at the sampling  point and
          purge the probe and  sample line by  repeatedly squeezing the  rubber suc-
          tion bulb until at least  7  air changes of the probe  and sample line have
          occurred.

-------
                                                                 Section No. 3-16.3
                                                                 Date June 30, 19B8
                                                                 Page 12

     7-   After  the probe  and the  sample  line are  completely purged,  leave the
          squeeze bulb in place, and open the inlet stopcock of the sampling flask
     8.   Leave  the  inlet  stopcock open until  the pressure  in the sampling flask.
          reaches the duct pressure.  This should take about 15 seconds.  Close the
          inlet stopcock.
     9.   Remove  the probe  from  the duct  and disconnect the glass tee  from the
          flask.
     10.  Tape the stopcocks  closed  and label  the  flask with the plant name, date,
          and sampling location, time,  and sampling personnel.
     11.  Immediately  after sampling,  determine  the flue  gas temperature  with a
          calibrated thermocouple, thermometer, or equivalent temperature measuring
          device, and  determine the static  pressure  of the duct  and the velocity
          over a period of time equal  to the predicted sample  run time following
          the procedures described in Section 3-1 in this Handbook.

      Purged Flask Procedure  - Collect  preliminary survey samples using the purged
flask procedure as follows:
     1.   Connect the  small  end  of  the sampling probe,  of sufficient  length to
          reach  the  centroid  of the  duct to  be sampled,  to the inlet stopcock of a
          precleaned glass sampling flask a sufficient length of Teflon tubing.
     2.   Connect the  rubber  suction bulb  to the other  stopcock with  a  piece of
          Teflon tubing or suitable flexible tubing.
     3-   Place  a plug of  glass wool in the  probe  inlet,  enlarged to approximately
          12-mm OD,   to serve as a filter to remove particulate matter.
     4.   Place  the  inlet  (filtered) end of  the  probe at  the centroid of  or no
          closer than 1 meter  to the duct wall.
     5.   Purge  the  probe, sample line,  and sample, flask by  repeatedly squeezing
          the rubber suction  bulb  until approximately 7 air  changes  of the system
          have occurred.
     6.   After  the  probe,  sample  line,  and  flask  are completely purged, close off
          the stopcock near the suction bulb,  and  then  close off the stopcock con-
          nected to the probe.
     7.   Remove  the probe from the duct,  and disconnect both the  probe  and the
          suction bulb from the flask.
     8.   Tape the stopcocks  closed  and label  the  flask  with the plant name,  date,
          and sampling location, time, and sampling personnel.
     9.   Immediately  after sampling,  determine  the flue gas temperature with a
          calibrated thermocouple,  thermometer, or equivalent temperature measuring
          device, and  determine the static pressure  of  the duct  and the velocity
          over a  period  of time equal  to the predicted sample  run time following
          the procedures described in Section 3-1 in this Handbook.

3.3.3   Flexible  Bag Procedure -  The  flexible bags used  to  collect  preliminary
survey samples must  be leak  checked  and demonstrated to be  free  of  contamination
following the  procedure described in  Subsection  3-2.2.   The preliminary  survey
sample collection using flexible bags can be conducted at a constant rate following
the  procedure  described in Subsection  4.3  for the  evacuated container  sampling
procedure, the direct  pump sampling procedure, or, in explosive  areas,  the explo-
sion  risk area  sampling  procedure.    The   flue  gas or  duct  velocity and  other
process parameters  should  be  determined  for designing the  final  sampling proced-
ures .

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 13

3.3.4  Adsorption  Tube Procedure - The  adsorption tubes used  to  collect the pre-
liminary survey sample(s) should contain  the  adsorbent  specified in Table B of the
Method Highlights  Section  for the target  analyte(s).   The sampling  time or total
sample volume  for  the  adsorption tube should be calculated based  on expected con-
centration^)  of  the volatile organic(s)  present  and the recommended  capacity of
the  adsorption media.    Refer to  the appropriate  reference  given  in  Table  B to
determine  the  recommended sample volume  taking into  consideration  the  amount of
adsorbent to be used.  For compounds not referenced in Table B,  use a reference for
a compound with  similar chemical characteristics.   If  the  target  analytes require
different adsorption media,  then it  is  recommended that preliminary survey samples
be  collected using  each  type of adsorbent.    In  the  case where  the  compound is
unlike any other  documented  compounds,  use two adsorption tubes connected in ser-
ies.   Once  a recommended  volume   is  established,  it  is   recommended   that  two
additional  samples  be  collected  with  sample  volumes  one  half  and  twice  the
recommended volume.      The procedure for collecting preliminary survey adsorption
tube samples is as follows:
     1.   Open the adsorption tube,  and  connect  the  primary  tube  section  (large
          section  of adsorbent)  to  the  sampling probe  using a minimum  length of
          Teflon tubing or other nonreactive tubing.
     2.   Connect  the  outlet  (backup section)  of the  tube  to the next  tube in
          series, if additional adsorption capacity is required.
     3.   Connect  the  outlet of  the  last tube to  the inlet of  the calibrated per-
          sonnel sampling pump using a sufficient length of tubing.
     4.   Insert the probe into  the  stack or duct and  turn on  the pump.   Maintain
          the  adsorption  tubes in a  vertical position during  sampling to prevent
          channeling.   Sample the gas stream  for  the time required  to obtain the
          optimal volume determined from the referenced method.
     5.   Immediately  after  sampling is  completed,  disconnect   the  tubes from the
          tubing and seal the tube  ends with teflon tape and plastic caps.  Label
          the  tubes  and store each  tube  in  a screw  cap  culture  tube or similar
          container to protect them during shipment.
     6.   Record  the  total  sampling time,  the sample  flow  rate,  the barometric
          pressure, and the  ambient temperature.

3-4  Preliminary Survey Sample Analysis and Interpretation

     With the  exception of the analysis of the glass  sampling flasks,  the analysis
of preliminary survey  samples should follow the procedures  described in Subsection
5.0.  The analysis of  the glass sampling flasks are described below  (see Subsection
3.4.2).   The  analysis  of  preliminary  survey  samples is  used  to optimize the
analytical procedures  and select  the most appropriate sampling  technique for final
sampling.    Using  Table C  the Method  Highlights  Section,  choose  appropriate GC
detector(s).   Based on the  sampling technique(s)  used  to collect the preliminary
sample,  choose a GC column  from the selections  listed in Table  D  of the Method
Highlights  Section;  the  technical  service department  of  column  manufacturers or
plant  laboratory personnel  may  also be  consulted  for additional  suggestions on
column type(s).    For  glass  flask  samples and  Tedlar or Mylar bag samples,  use
calibration  gas  cylinders  or calibration  standards  prepared  in Tedlar bags.   For
adsorption   tube   samples,   prepare   the   calibration  standards directly  in  the
desorption liquid(s) or on adsorption tube material(s) used to collect  the samples.

3.4.1  Calibration Standards for Preliminary  Survey Samples - Prepare a minimum of
three calibration  standards  for  each compound  of  interest.   The  standards should

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 14

cover a  linear  range for the particular GC detector,  with  the lowest standard and
the highest standard bracketing and a midrange standard approximating the expected
sample concentration.   To  estimate the sample concentration,  prepare a preliminary
survey sample and  perform  a single analysis  of  the  sample  following the procedure
in  the  appropriate  subsections  below. During  this  sample analysis,  determine if
adequate resolution  has been achieved for each peak with a peak area greater than
5#  of the  total chromatographic peak  area  (excluding  the  desorption solvent peak)
using one of the procedures  described in  Subsection  3-4.1.   Adequate resolution of
sample peaks  will  only be  necessary in the chromatographic region(s)  where the
target compound(s)  are  expected to elute.  The GC analysis conditions and/or column
can be changed  to  achieve adequate  resolution.   The  use of  two different columns
may be necessary to ensure accurate identification of the  gases.   For analysis of
more  than  one  target  compound in  very complex  sample matrices,  more  than one
analysis using  different GC conditions and/or columns may be  required to achieve
adequate resolution  for all target compounds.
     For analysis of flask samples  or bag samples,  (1) use  cylinder gases directly
(if available)  or  by  dilution  following the  procedures  described  in Subsection
5.1.1 and  5.1.2,  respectively,  or  (2)  prepare standards in  Tedlar bags following
the procedure described in Subsection 5.1-3 for gaseous materials or the procedure
described in  Subsection '5-1.4  for liquid materials.   For analysis  of adsorption
tube samples, prepare  calibration  standards  following the procedure  described in
Subsection 5-1.6.    Data forms  should be used for recording  calibration standard
preparation and analysis data (see  Figures 5-4,  5-6,  5-8,  and 5«9) and preliminary
survey sample analysis data  (see Figure 5-1)•
     The gaseous calibration standards for bag samples must be injected into the GC
using a  gas  sampling valve equipped  with a  stainless steel or  Teflon sample loop
following the procedures described in Subsection 5-1  appropriate for the particular
type of  gaseous standard used.    Liquid calibration  standards  for adsorption tube
analysis must be injected into a heated sample injection port following . the proced-
ure described in Subsection  5-1.6.   The gaseous standards  for  glass flask samples
can be   injected  into  the  GC  using  either   a gas  sample  valve,  following the
appropriate procedure in Subsection 5-1 for  the particular gaseous standard used,
or  a  heated  injection  port  using  a gas  tight syringe  following  the  procedure
described below; the same  injection procedure used for  the standards  must be used
for the  flask samples.
     The procedure  for  injecting gaseous calibration  standards using  a  gas tight
syringe  is as follows:
     1.    Attach a GC septum to a. piece of Teflon tubing and  attach the tubing to
          the outlet  of the  calibration gas cylinder  regulators or the Tedlar bags
          containing the calibration gases.
     2.    Insert the needle of the syringe through the septum, and repeatedly purge
          the syringe by repeatedly filling and emptying the syringe 7 times.
     3.    After  purging  the  syringe,  fill .the syringe past the  mark corresponding
          to the desired amount to be injected, and withdraw  the  syringe from the
          septum.    Stick the needle  into a  rubber  stopper  or  a  thick  septum to
          prevent dilution of the standard by  ambient air.
     4.    Immediately before injecting the standard,  remove  the  needle  from the
          stopper or  septum,  adjust the syringe to the  desired  volume,  and inject
          the standard  into  the heated injection port  on the  GC.   Note the time of
          injection on the strip chart and/or  actuate the electronic integrator.
     5-    Repeat the  injection  of the standard until  the peak  areas from consecu-
          tive injections agree  within 5% of their average  value.

-------
                                                                 Section No.  3-16.3
                                                                 Date June 30, 1988
                                                                 Page 15

     Perform initial tests using the calibration standards to determine the optimum
GC conditions to minimize  analysis  time  while  still maintaining sufficient resolu-
tion.  Sufficient resolution can be determined following the procedure described by
Knoll42 or  in EPA  Method 625^* where  the baseline  to  valley height  between two
adjacent peaks must be less than 25% of the sum of the two peak heights (see Figure
3-3).
     Analyze  the  calibration  standards,  starting  with  the  lowest  concentrations
first.  Repeat  each standard analysis until two  consecutive  injections give indi-
vidual area  counts within 5%  of  their average.   Multiply the average of  the re-
sponse for the two acceptable consecutive injections of each standard by the detec-
tor attenuation to  determine the calibration area value  for each standard.   Record
the  retention  time for each compound  and the  calibration  area  for  each standard.
Record  the  detector  settings,   the  recorder/integrator  attenuation  for  each
standard,  the chart speed, the GC temperature settings, the column parameters  (type
and  length) ,  and  the carrier gas  flow  rate.   Plot the concentration  of  the stan-
dards on  the abscissa (x-axis) and  the  calibration area  for  each standard  on the
ordinate.    Perform a regression analysis,  and  draw the least squares  line  on the
plot.  It is  recommended,  but  not  required for preliminary survey sample analysis,
that  the  validity  of the  calibration  curve be checked using  the audit procedures
described in Section 8.0.  The audit sample may be analyzed at this time in lieu of
analysis  during   the  final  sample analysis  with  the   prior  approval  of  the
Administrator.
     If positive identification of a target compound  cannot  be  made by comparison
of the compound retention  time  to  the  retention time of  one of the standards, then
use of a different type of column may be helpful.  If positive identification still
cannot be achieved,  then  GC/mass spectrometry  (GC/MS)  or GC/infrared (GC/IR) tech-
niques should  be used, with  GC/MS recommended.   In addition, any  compounds, not
identified as target compounds, with peak areas greater  than  5#  of the total chro-
matographable  peak area  (excluding the  solvent  peak area for  adsorption  tubes)
should be identified by comparison to known standards or by using GC/MS.

3-4.2  Glass Flask Preliminary Sample Analysis - Since glass sampling flasks are
only used for  preliminary survey  samples, the  analysis of the flasks  is  described
in this section.  Glass sampling flasks require some pressurization prior to analy-
sis to withdraw the sample.
     Using  the  ideal gas  law,  the amount of  dilution of  the sample  that results
from pressurization can be estimated with enough accuracy to permit  interpretation
of the preliminary survey sample  results.  The procedure  for pressurizing a flask
is as follows:
     1.   Note if any condensation has collected in the flask.  If it has, heat the
          flask to  the flue gas or  duct  temperature with  an oven, heating tape, or
        •  a  heat  lamp.   Note: The  pressurization of sealed glass containers by
          heating  is  an  inherently  hazardous   process.    The  use of  a protective
          shield  to  protect   personnel   from   flying  glass  in  the event  of  an
          explosion  is highly  recommended.    In addition,  the  flask should  be
          wrapped in cloth or other cushioning media during these operations.
     2.   Connect one  end  of  the  flask to a mercury manometer,  open the  stopcock,
          and determine the initial  pressure of the  flask  (P.).   Record Pi  and the
          initial absolute flask temperature (T. )  in °R or °K.
     3.   Connect the other end of the flask to a source of hydrocarbon-free nitro-
          gen or  air,  and open the stopcock.   Slowly pressurize  the flask  to  a
          maximum of 15 psig,  and close the stopcock.  Determine the  final pressure
          of the flask  (Pf)  and the final absolute temperature of the  flask (T. ) .

-------
                                                                I
p
r
                                         p
                                         r
                              V
                                                             O 0>
                                                             CD CD

                                                             ct O
                                                             CD ct
                                                               H-

                                                             t-i O


                                                            '§*
                                                                                                 O
                                                                                                 -  00
Figure 3.3.  Diagram showing EPA Method 625 criterion  for  adequate resolution of

             overlapping compounds with similar mass spectra.
                                                              OQ •
                                                              O3OJ

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 17

          Note:  The  pressurization  of sealed  glass containers  is  an  inherently
          hazardous process.   The  use of a protective shield  to  protect personnel
          from flying glass in the event of an explosion is highly recommended.  In
          addition, the flask  should  be wrapped in  cloth or other cushioning media
          during these operations.
     4.   Calculate the first dilution factor (D1)  using the following formula:

                                        Pf  xT,
                                   D,  = 	                      Equation 3~2
                                        Tf  x  PI

     5.   Allow the flask  to equilibrate  for  10 minutes.   Note if any condensation
          has formed. If condensation has formed in the flask and the flask did not
          initially required heating, heat the flask to a temperature sufficient to
          vaporize the condensate.   If the condensate cannot be vaporized or if the
          flask was  already heated  and  more condensate  formed  during pressuriza-
          tion,  the sample cannot be analyzed accurately.
     6.   Close the stopcocks  and disconnect the manometer and dilution gas.
     Analyze the contents of a pressurized flask using a sample introduced into the
GC via a gas sampling valve by the following procedure:
     1.   Connect  the  sample  flask  to the injection valve  with the  valve in the
          load position.
     2.   Open  the stopcock connected  to the valve, and allow the  gas  sample to
          flow  through  the sample  loop  at 100  ml/min  for  30  seconds (determined
          with  a  rotameter connected  to  the outlet  of  the sample  loop)  or purge
          with 5 times  the  sample  loop volume,  whichever is less.   Close the stop-
          cock,  and allow the  sample loop to return to ambient pressure.
     3.   Actuate  the  sample valve to inject  the  sample and  record the injection
          time.
     4.   Examine  the  chromatogram and determine  if adequate resolution  has been
          achieved  between individual  target  compound  peaks  and  between target
          compound peaks  and  any  interfering compound  peak with an  area greater
          than 5$  of the  total area of all peaks (excluding the desorption solvent
          peak)  using the procedure described in Subsection 3-4-1-
     5.   Determine  the  retention  time for each peak by dividing  the distance of
          the peak maximum from the injection point by the chart speed.
     6.   Repeat the analysis,  and determine the peak area  and  retention time for
          each target compound identified during the second analysis.  Although not
          required for  the preliminary survey sample analysis, the  peak areas for
          each target compound from consecutive injections  should  agree  within 5%
          of the average peak area.  The retention times between the two injections
          should agree  within  0.5  seconds or  \%  of  the adjusted  retention time
          (compound retention  time minus  the  time  of elution of unretained peaks),
          whichever is greater.
     Analyze the contents of a pressurized flask using a sample introduced into the
GC via a gas tight  syringe and a heated injection  port sample by  the  following
procedure:
     1.   Attach a GC  septum to one  of  the stopcocks on the glass  flask.   (Note:
          Glass  sampling flasks can be purchased with an integral septum port.)
     2    Insert the needle of the syringe  through  the  septum,  and  purge  the sy-
          ringe  by repeatedly filling and emptying the syringe 7 times.
     3.   After  purging the  syringe,  fill the syringe past  the  mark corresponding
          to the desired  amount  to be injected, and withdraw  the  syringe  from the

-------
                                                                 Section No.  3.16.3
                                                                 Date June  30,  1988
                                                                 Page 18

          septum.   Stick  the  needle into  a rubber  stopper  or a  thick  septum  to
          prevent dilution of the standard by ambient air.
     4.   Immediately  before  injecting the  sample,  remove  the  needle  from  the
          stopper  or septum, adjust  the syringe  down  to the  desired  volume,  and
          inject  the sample into the  heated injection port  on the GC.   Note  the
          injection  time  on  the strip  chart and/or actuate the electronic  integra-
          tor.
     5-   Determine  the  retention time  for  each peak by  dividing  the  distance  of
          the peak maximum from the injection point by the chart speed.
     6.   Repeat  the analysis,  and  determine the retention times for each peak  for
          the second analysis.  The retention times of successive injections  should
          agree  within  0.5  seconds  or within 1%  of  the  mean  of the   adjusted
          retention times, whichever is greater.

3.4.3  Analysis of Preliminary Survey Bag Samples  - Follow the procedures described
in Subsection  5-3-1  for the analysis of bag samples.  To assess  the  stability  of
the gas sample in Tedlar bags, perform a second analysis after a time period  equal-
ling the  period  between  sample collection and  the first  analysis.   If the concen-
tration of the sample  collected in  a Tedlar bag decreases by more than 10% between
the first and  second analysis,  then an  accepted sampling method other  than  Tedlar
bags should be considered..
     Perform a retention check on the bag sample by successively evacuating the  bag
and refilling it  with  hydrocarbon-free  air  or nitrogen one or more times.   Analyze
the bag contents  for  the  target compound(s),  allow the  gas  to  sit  in  the  bag
overnight, and  reanalyze bag contents  for  the target compound(s).  If any  target
compound is detected in  the  bag at  a concentration greater than 5/» of the original
concentration, then  an accepted sampling method other than Tedlar  bags  should  be
considered.

3.4.5  Analysis  of Preliminary  Survey Adsorption Tube Samples - Follow the proced-
ures described in Subsection 5-3-4  for  the  analysis of adsorption tube samples.  A
minimum desorption efficiency  of 50# must be obtained.   If  50% desorption effici-
ency cannot be achieved  using  the referenced procedures  from  Table. B in the  Method
Highlights Section,  then  try  longer  desorption  times,  more  vigorous desorption
techniques and/or other desorption  solvents.   If  50% desorption  efficiency still
cannot  be accomplished,  then  an accepted  sampling  method  other  than adsorption
tubes should be considered.

3.4.6   Interpretation  of Preliminary Survey Eesults  - To  select  the most suitable
sampling and analytical method for the final field test,  the results of the prelim-
inary survey must be properly interpreted.   The major points  to  consider are  (1)
the sampling location,  (2) the parameters of the process being tested,  (3)  the flue
gas moisture and temperature and the flue or duct  static pressure,  (4)  stability  of
the gas sample in bags,  (5)  the desorption  efficiency of the  target compounds from
adsorption tubes, and  (6)  the  resolving capability, precision,  accuracy,  and speed
of the  GC analysis.   Thus,  flue gas  or  duct  parameters and  components  present
determine which sampling and analytical methodologies will be the most appropriate.

     Sampling Location  - The hazards  associated  with  the sampling location will .
influence the type of  sampling methodology  which  can be used.   In  explosion risk
areas  where use  of pumps,  heated probes, or a  GC  with a  flame  ionization  detector
(FID)  would be prohibited,  the  explosion risk area sampling procedure  can  be used
safely.   Close attention must be paid to maintaining the  proper sampling rate when

-------
                                                                 Section No. 3-16.3
                                                                 Date June 30, 1988
                                                                 Page 19

using  the  evacuated canister as  a vacuum source.   It may be possible  at certain
hazardous  locations,  to locate the  collection  device (bag or adsorption tube)  in
the explosion risk  area near the source and operate  the  sampling  pump  a safe dis-
tance  away.   Also,  intrinsically safe personnel  sampling  pumps could  be  used  in
certain  hazardous  areas  provided the  required sampling  rate  can  be  maintained.
Hazardous  sampling  locations  may not be suitable  for direct  or  dilution interface
sampling when using an electrically-heated probe and sample line.
     Other physical  factors concerning  the  sampling location will  also influence
which  sampling  method is most suitable.   These factors will be site-specific and
are beyond the scope of this Handbook.

     Process  Parameters  -  The  particular process  parameters  pertaining to  the
generation  of  the  organic  emissions  and the  effect the  operation  has on  the
emission levels will  influence which sampling  technique will  be  most suitable.   In
the case of  a continuous process  where emission levels are  constant,  each of the
Method  18   sampling  techniques  should  be suitable  with  regard  to  the  process
parameters.  For processes  operating 'in a  batch or cyclic mode,  the bag or adsorp-
tion  tube  integrated  sampling techniques  may  be more  suitable  compared to  the
interface  techniques where grab samples are analyzed.

     Flue  Gas or  Duct Conditions  -  The flue  gas  or  duct  moisture and  temperature
will  have  a  major  influence on  selecting the  most suitable sampling technique.
High moisture will affect both bag samples and adsorption tube samples.  For situa-
tions  where  moisture may be a problem,  the  interface techniques  are  recommended
provided the 5% criteria for consecutive injections, described in Subsections 5-3-2
and 5-3-^ •  can be met.  Condensation in bag samples may result in the target organ-
ic  compounds being  absorbed into  the condensate, or,  at extremely  high concen-
trations,  being the condensate itself.  The  heated bag sampling  technique may  be
suitable provided  on-site analysis  is  conducted when it is  not practical to keep
the bags heated until analysis at  the base laboratory.  Condensation  may also  be
avoided by using a  diluted  bag sample collected by prefilling the bag with a known
quantity of hydrocarbon-free  air  or nitrogen  prior to  sampling  and  accurately
metering the gas sample into the bag during sampling.
     Moisture reduces the adsorptive capacity  of certain  types  of adsorbents (pri-
marily  charcoal).    For sampling  with adsorption tubes  at  sources with moisture
above  3$.  a silica  gel tube may  be inserted  in  front of the  primary adsorption
tube;  otherwise,  two or more adsorption tubes  connected  in series should be used.
The  first  tube  becomes  a  sacrificial  tube  and  should  be  positioned  vertically
during sampling.  A disadvantage of  this approach  is  that the additional tubes will
also  require analysis.    Alternatively, a moisture knock-out jar can  be  used  in
front  of  the adsorption  tube.   As  varying amounts  of the organic  emissions will
also condense (the  amount of  each organic  removed from the gas  stream will largely
depend on  the individual compound's volatility and solubility characteristics), the
liquid collected must be retained for  analysis.   Accurate  quantitation of various
organics in  the condensed  liquid(s)  may  involve several  steps and  is generally
problematic.
     The flue gas temperature may also dictate which sampling technique can be used
due to limitations of the sampling equipment.

     Bag Sample Stability and Target  Compound  Retention  - IF on-site  analysis  of
bag samples is not feasible and the samples are returned to the  base laboratory for
analysis,  then  the  stability of  the gas  sample in the bag  will  be a  factor  and
should be determined.  While the  stability of organics  in  bags has been demonstra-

-------
                                                                 Section No. 3-16.3
                                                                 Date June 30, 1988
                                                                 Page 20

ted in numerous laboratory evaluations, an actual source sample could contain other
unknown  components  which may affect  sample  stability.   If  the preliminary survey
sample analysis indicates that the gas sample is not stable, then one of the alter-
native sampling  procedures should be used.   The  check should be  conducted at an
elevated temperature if the bag is to be heated during sampling.
     The retention of  the  target  compounds by the  Tedlar bag should also be check-
ed.   This  check will  indicate  any  sample  loss not  determined by  the stability
check.   If the  retention  of a target compound  by the Tedlar  bag  is unacceptably
high, then  the bag sampling  technique is  not suitable  for that target compound and
one  of  the alternative  sampling procedures  should  be used.   Heating  of  the bag
during sampling and analysis may reduce the retention.

     Adsorption  Tube  Desorption  Efficiency - The  desorption efficiency determined
for  the  adsorption  tubes must be >50#.    If >50#  desorption  efficiency cannot be
achieved with  the referenced procedure,  then more vigorous desorption techniques
and/or solvents  should be  evaluated.   The desorption efficiency,  as  determined by
the procedures described in  Subsection 5-1.6, will not indicate  if the gas sample
matrix  will  affect   the  desorption  of  the  target   compounds.    If  acceptable
desorption  efficiency cannot be  achieved,  then one of  the  alternative  sampling
procedures  should be  used.   Also, the adsorption  efficiency must  be  greater than
90%. The breakthrough volume must not be exceeded.

     Calibration Standards and GC Analysis - The availability of calibration stan-
dards may dictate which sampling technique can be used.  The  GC analysis  may also
dictate which sampling technique will be the most suitable.   For accurate analysis,
adequate  resolution  must  be achieved  between  target  compounds  and  between  any
interfering  compounds  and  target compounds.   During  preliminary survey  sample
analysis, acceptable  resolution  may  not be  achievable  on  a gas  sample but  may be
accomplished with the adsorption tube sample,  or  vice versa.   Thus,  the  sampling
technique which gives  acceptable  resolution  during sample  analysis  must be select-
ed.  In  some situations where analysis of more  than one target compound is requir-
ed, two  or  more  analyses of the  same sample under different  GC conditions and/or
with different columns may be necessary to achieve  adequate resolution.
     Acceptable  accuracy,  as demonstrated by audit  sample  analysis,  must  also be
achieved for sample analysis by either gas or liquid injection.  Again the sampling
technique that gives  acceptable  accuracy during sample analysis  must  be selected.
The sampling technique that gives acceptable precision,  as demonstrated by consecu-
tive replicate injections,  must be selected.  Minimizing  the analysis  time is par-
ticularly important for  the  interface techniques.   As  discussed above, the preci-
sion limits may be hard to achieve with the interface techniques with a long analy-
sis time under variable or cyclic emission conditions.

3-5  Apparatus Check and Calibration

     Figure 3-^ summarizes the pretest apparatus checks and calibration and can be
used  as  a  pretest  operations checklist.    Figure  3-5  can  serve as  an equipment
packing list and status report form.

3-5•!  Probe - If a heated probe is  required for  the  selected sampling procedure,
then the probe's  heating system should be checked to see that it is operating prop-
erly.   The  probe should be  cleaned  internally  by  brushing first  with tap  water,
then with deionized distilled water,  and  finally with acetone.   Allow  the  probe to
air dry,  then the probe should be heated and purged with air or nitrogen.  The

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 21

Date 	  Calibrated by 	

Check Sampling Technique To Be Used

Direct Bag 	,  Dilution Bag 	,  Direct Interface 	,  Dilution Interface 	,
or Adsorption Tube 	.

Sampling Checks (Check only applicable methods)

Velocity and Water Vapor Content

Pitot tube dimension specifications checked?	yes 	no  (specification of Method
2, Handbook Section 3.1)

Differential pressure gauge pretest calibration acceptable? 	yes 	no 	N/A
(specifications of Method 2, Handbook Section 3-2)

Stack temperature sensor calibrated against a reference thermometer?* 	yes 	no
(within 5°F of reference thermometer)

Barometer pretest field barometer reading correct?  	yes 	no (within 2.5 mm
(0.1 in.) Hg of the mercury-in-glass barometer)

Wet bulb/dry bulb thermometers accuracy acceptable?  	yes 	no (within 1°F of
true value, manufacturer's specifications)

Method 4 sampling equipment acceptable?* 	yes 	no (Handbook Section 3.3, PRE
TEST SAMPLING CHECKS, Method 4, Figure 2.5)

Direct Bag

Pretest calibration of flowmeter acceptable?  	yes 	no (within 10 percent of
0.5 liter/min for single check)

For heated box system, pretest calibration of the temperature sensor in the box is
acceptable?  	yes 	no 	N/A (within 5 percent of reference value at
temperature of expected use)

Dilution Bag

Pretest calibration of flowmeter acceptable?  	yes 	no (within 3 percent of wet
test meter)

Pretest calibration factor of dry gas meter acceptable?   	yes 	no (within 2
percent of wet test meter)  	


*Most significant items/parameters to be checked.

(Continued)
                      Figure 3.4.  Pretest sampling checks.

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 22

Figure 3.4  (Continued)

Direct Interface and Dilution Interface

Pretest calibration of stack temperature sensor acceptable?*  	yes 	no (within
2°F of reference value)

Pretest calibration of probe and heated sample line temperature sensor acceptable?*
	yes 	no (within 2°F of reference value)

For dilution interface only, pretest calibration of dilution system acceptable?*
	yes 	no (within 10 percent of expected dilution factor)

Pretest calibration of gas chromatograph acceptable?*	yes 	no (specifications
shown in POSTSAMPLING OPERATIONS CHECKLIST, Figure 5.10)

Adsorption Tubes

Pretest calibration of limiting orifice acceptable?*	yes 	no (compared to
bubble meter)  .                    ,,.-.,.

*Most significant items/parameters to be checked.

-------
                                                                 Section No. 3-16.3
                                                                 Date June 30, 1988
                                                                 Page 23
Apparatus check
Moisture Determination
N Bulb/D Bulb
Checked
Barometer
Calibrated*
Method 4
Probe, heated &
leak checked
Impingers
Meter system
calibrated*
Velocity Determination
Pi tot Tube
Number
Length
Pressure Gauge
Manometer
Other
Stack Thermometer
Calibrated 	
Bag Sampling
Probe Liner
S steel
Glass
Teflon tube
Length
Meter System
Flowmeter*
Pump
Evacuated can
Charcoal tube
Sample line
Tedlar Bags
Number
Blank checked
Heated Box
Number
Heat checked
Acceptable
Yes



No



Quantity
Required



Ready
Yes



No



Loaded and Packed
Yes



No



*Most significant items/parameters to be checked.

                        Figure 3-5'  Pretest preparations.

-------
                                                                  Section No.  3.16..3
                                                                  Date June 30,  1988
                                                                  Page 24
Figure 3-5  (Continued)
Apparatus check
Bag Sampling (continued) •
Dilution
Inert gas
Meter
Gas Chromatograph
On- site
(check below)
N/A

Direct or Dilution
Interface
Probe Liner
Glass
S steel
Teflon
Heated Line
Length
Heat checked
Temperature Sensors
Stack
Probe
Calibrated*
Sample Pump
Dilution System
Dilution pumps
Flowmeters
Dilution gas
Heated box
Dilution factor
checked*
Gas Chromatograph
(shown below)
Adsorption Tube
Probe
Heated
Checked
Nonheated
Glass
S steel
Filter
Acceptable .
Yes
i
\
:
\

NO ;
f
>
i
t
I
Quantity
Required



Ready
Yes



No



Loaded and Packed
Yes



No



*Most .significant items/parameters to be checked.

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 25
Figure 3-5  (Continued)
Apparatus Check
Adsorption Tube (continued)
Sample Line
Type
Length
Checked*
Pump and Meters
Pump
Orifice
Calibrated*
Rotameter
Calibrated*
Timer
Adsorption Tubes
Time

Gas Chromatograph
Cylinder Standards
Ana lute
PPM
PPM
Regulators
Bags
Size
Dilution system
Calibrated*
Diluent gas
Bag Standards
Analyte
Syringes
Imptnger/hot
plate assembly
Gas meter
N2 gas
Regulator
Bags
Size
Other Gases
Fuel
Carrier
Zero
Co lumns
Type

Acceptable
Yes


No


Quantity
Required


Ready
Yes


No


Loaded and Packed
Yes


No


*Most significant items/parameters to be checked.

-------
Figure 3-5  (Continued)
                                                                 Section No. 3.16-3
                                                                 Date June 30, 1988
                                                                 Page 26
Apparatus Check
Gas Chromatograph
Type
Temp/con oven

Bulk Samples
Bottles
Type
Size
Acceptable
Yes


No


Quantity
Required


Ready
Yes


No


Loaded and Packed
Yes


No


 *Most significant items/parameters to be checked.

-------
                                                                 Section No. 3-16.3
                                                                 Date June 30, 1988
                                                                 Page 27

probe should be sealed at  the  inlet  end  and checked for leaks by applying a vacuum
of 10 in. H20.   See Subsection 1.0 for the probe  leak  check  procedure.   The probe
is considered  leakfree  under these conditions  if  no loss of vacuum  is  seen after
one minute.  • Any leaks should  be  corrected or the  probe should be rejected. If the
probe has an external sheath,  the integrity of the seal between the sheath and the
probe liner should be checked to ensure ambient air does not dilute the gas sample.

3.5.2   Teflon  Tubing -  Prepare enough sections of tubing for connecting the probe
to bag or tube samples .

3.5.3   Quick Connects or  Equivalent - The quick  connects,  or  their equivalents,
should  be new or  cleaned  according to  the manufacturer's recommendations.   Leak
check the quick connects as described in Subsection 1.0.

3.5.U  Barometer - The field barometer should be compared with the mercury-in-glass
barometer or  with  a National  Weather  Service Station reading prior  to  each field
test.
3-5-5   Wet  Bulb/Dry Bulb Thermometers - For  sources  with stack temperatures below
59° C  where  wet  bulb/dry bulb  thermometers will  be  used  to determine  stack gas
moisture  content,  the  thermometers  should be  compared with  the mercury-in-glass
thermometer at room temperature prior to each field trip.

3-5-6   Method 4 Equipment  - Where  Method 4 will  be used to  determine  stack gas
moisture  content,   prepare  the  equipment  for  sampling  following  the  procedures
described in Section 3-3-3 of this Handbook.

3-5-7  S-type Pitot Tube and Differential Pressure Gauge - Prepare the S-type pitot
tube and  the  differential pressure gauge  for sampling following the procedures de-
scribed in Section 3-1-3 of this Handbook.

3-5-8   Sampling  Pump  - Check the  sampling  pump  for delivery  rate and leaks before
going  to  the field  as follows:   Attach  a 0 to  5 liter/minute rotameter  to the
outlet  of the pump and turn on the  pump.   Check  the  flow rate  indicated  by the
rotameter.  Reject  or  repair  the pump if  the flow rate is not at least 1 liter/mi-
nute.  If the flow is adequate, then conduct a leak check by plugging the inlet of
the pump.   If the  pump is leak free  then  the rotameter should eventually indicate
no flow.  Repair or replace the pump if a leak is indicated.

3-5-9  Tedlar Bags - Prepare new Tedlar bags for sampling by leak checking the bags
before  going  to  the field.    The bags should also  be checked  for contamination by
filling with  hydrocarbon- free air  or  nitrogen during  the leak check.  The bags are
checked as  follows:   Connect  a water manometer,  or  equivalent, using a tee con-
nector between the  check valve quick connect on  the  bag and  a pressure source (or
hydrocarbon- free air  or nitrogen  for  conducting the contamination  check) .   Pres-
surize  the  bag  to  5 to 10  cm (2  to  4  in.) H20 and  disconnect the quick connect.
Loss of pressure over  a 10  minute period  indicates  a  leak.   Alternatively, leave
the bag pressurized overnight; a deflated  bag the following day is indicative of a
leak.   Reject or repair  any bags  with leaks.   After  the  hydrocarbon- free  air or
nitrogen has  remained  in the  bag for 24  hours,  analyze the bag contents using a GC
with a  flame  ionization detector on  the most sensitive  setting.  The bag should be
rejected if any organic compounds  are detected that may interfere with the analysis
of any of the target compound(s).

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 28

3-5-10   Etgld Leak-Proof  Containers -  The rigid containers  used to  contain the
Tedlar  bags  during  sampling should  be  checked  for  leaks prior  to going  to the
field.   The  container  should be leak checked with the  bag in place  as follows:
Connect a water manometer,  or equivalent,  using a tee  connector between a pressure
source  and the  container outlet.  Pressurize  the container to 5  to  10 cm (2 to 4
in.) Hg.  Any loss of pressure after 10 minutes indicates a leak.  Reject or repair
the rigid container if a leak is indicated.

3.5-H  Direct Pump Sampling System - If the direct pump sampling system is select-
ed,  then  the system  should be  assembled  and  leak checked  prior  to going  to the
field as follows:  Assemble the  system  (see Figure 4.5).   Attach a vacuum line and
a  rotameter  to the  inlet  quick connect.   Plug  the  probe  inlet  and  turn  on the
vacuum  pump.   If  the  system is leakfree  up to  the pump, the rotameter should even-
tually  indicate no flow.  An alternate procedure to leak check the system up to the
male inlet check  valve  quick connect  is  as follows:   Connect a water manometer, or
equivalent, using a  tee connector between  a pressure  source and  the inlet end of
the  probe.   Pressurize  the system  to 5  to 10 cm  (2  to 4  in.)  Hg.    Any loss of
pressure after 30 seconds  indicates  a leak.   Reject or repair the sampling system
if  a leak  is indicated.   Check to see  if the pump is  contaminating the sampling
system  by filling a  second contamination-free  Tedlar  bag with hydrocarbon-free air
or  nitrogen,  and  with the  system assembled pull  the hydrocarbon-free air or nitr-
ogen from the second  Tedlar bag  into  the first Tedlar  bag using the pump.  Analyze
the  first  bag contents  using a GC with a flame  ionization detector  on  the most
sensitive setting.   The pump should be rejected  or repaired,  cleaned,  and checked
again if any organic compounds are detected that may interfere with the analysis of
any of  the target compound(s).

3-5-12  Needle Valve and Rotameter - Prior to each field trip or at sign of erratic
behavior, the flow control valve and the rotameter should  be  cleaned according to
the maintenance procedure recommended by the manufacturer.

3.5-13  Teflon Probe  -  For bag  sampling in an explosion risk area,  prepare  a new
Teflon  probe  or  clean  a used Teflon probe following  the procedure described  in
Subsection 3-5-1-   Leak check the Teflon probe  as follows: Attach a mercury manome-
ter, with a tee connector,  and a vacuum  pump to the outlet of the  probe.  Plug the
inlet end of the  probe  and apply a vacuum of  10 in.  H20.  The probe  is considered
leak free under  these conditions if  no  loss  of  vacuum is seen after  one minute.
Any leaks should be corrected or the probe should be rejected.

3-5-14   Explosion Risk  Area Sampling System  - The explosion risk  area sampling
system  should be  leak checked as follows:   Evacuate the  steel drum.   Assemble the
system  (see Figure 4.6), with the pinch clamp open, the sample bag leak checked and
evacuated,  and directional needle valve closed.   Attach a mercury  manometer to the
inlet of the  Teflon probe.   Open the needle valve.  The rotameter should eventually
indicate no flow.  Once there is no flow,  note the manometer  reading.   The system
is considered leak free  under these  conditions if no loss  of  vacuum  is seen after
one minute.   Any leaks should be corrected or the system should be rejected.  It is
recommended that  an  explosion-proof  pump be used  in  the explosion risk  area or a
regular pump be used outside the risk area.   Follow the  procedures  described for
these pumps.

3-5-15  Heated Bag Sample Container and Sample  Lines -  If other modified bag sampl-
ing  techniques are selected due to  condensation observed  during  sampling,  heated

-------
                                                                 Section No. 3-16.3
                                                                 Date June 30, 1988
                                                                 Page 29

bag sample  containers  and sample lines  will  be required.  The  heating systems of
this equipment should  be  checked prior to going to the  field  to see that they are
operating properly.   The  sample lines  should  be  cleaned following the procedure
described for  the  sampling probe in Subsection 3-5-1-   The  heated sampling system
should  be assembled  and   leak  checked  prior to going  to  the  field  as follows:
Assemble  the system  (see  Figure  4.5).   Attach a vacuum pump and a rotameter to the
inlet quick  connect.   Plug  the  probe inlet  and turn  on the vacuum pump.   If the
system  is leak free,  the  rotameter should eventually  indicate no  flow.  An alter-
nate procedure to  leak check the system up to  the  female outlet check valve quick
connect on the bag container is  as  follows:   Connect  a water manometer, or equiva-
lent, using  a tee connector between  a  pressure  source  and  the inlet  end  of the
probe.  Pressurize the system to 5  to 10 cm (2 to  4 in.) Hg.  Any loss of pressure
after 30  seconds indicates a leak.   Reject or repair  the sampling system if a leak
is indicated.

3.5-16   Direct Interface  Sampling System  -   The  heating system of  the sampling
probe should be checked  prior  to  going  to  the  field  if heating  is  required to
maintain  the gas sample above the duct temperature and/or to prevent condensation.
The probe should also be cleaned and leak checked following the procedures describ-
ed in Subsection 3-5-1-   If the  probe has an external sheath,  the integrity of the
seal between the sheath and  the probe  liner should be checked to ensure ambient air
does not  dilute  the  gas  sample.   The sample line should  be  cleaned following the
procedure described for the  sampling probe in Subsection 3-5-1-  The heating system
of the  sample  line should be checked  before  going to the field to  see that it is
operating properly.   The  direct interface sampling system should  be assembled and
leak checked prior to going to the  field as follows:   Assemble the  system (see
Figure  4.5).   Switch the  gas sampling valve  to the inject position,  and plug the
outlet  from  the  sample valve.   Connect  a water manometer, or equivalent,  using a
tee connector between a pressure source and the inlet end of the probe.  Pressurize
the system to  5  to 10 cm  (2 to  4  in.) Hg.  Any loss  of pressure  after 30 seconds
indicates a leak.  Reject or repair the sampling system if a leak is indicated.

3-5-17   Dilution Interface  Sampling  System  - The equipment required  for dilution
interface sampling is  the  same as  required for direct interface  sampling, with the
addition of a heated dilution system  and a larger  heated sample  pump.   The heating
systems should be  checked to see that they are operating properly.   Prior to each
field trip or at sign of erratic behavior, all flowmeters should be cleaned accord-
ing to  the  maintenance procedure recommended by the  manufacturer.   The flowmeters
should  also  be calibrated following the procedures described  in Subsection 2.1.3-
The dilution interface sampling system should also be checked for leaks as follows:
Assemble  the  system  (see  Figure 4.6).   Connect a water  manometer,  or equivalent,
using a tee connector between a pressure source and  the inlet  end of the probe.
Plug the  three outlet vents to  the charcoal adsorbers  and  the  outlet  of  the two
flowmeters.    Pressurize  the system to 5 to  10 cm  (2 to  4  in.)  Hg.   Any  loss of
pressure after 30 seconds  indicates a leak.   Reject or repair the system if a leak
is indicated.  It is advisable to verify the operation of the dilution system prior
to going  to  the  field following the procedures described  in Subsections 4.3-7 and
5-3-3-

3.5.18   Gas Chromatography System -  Refer  to  Table C  in  the Method Highlights
Section to ensure that the proper detector has been selected for the target organic
compounds.  Prior to taking the  gas chromatography  system to the field, check that
all systems  are  operating properly.    Consult the operator's manual  for procedures

-------
                                                                 Section No. 3-16.3
                                                                 Date June 30, 19*88
                                                                 Page 30

to verify that the equipment is operating properly.  Check to see that all cylinder
gas regulators, connections, and tubing are functioning properly and are leak free.
This is  particularly important when using  hydrogen and oxygen.   Consult  with the
regulator manufacturer for procedures on checking pressure regulators.  Connections
and tubing can be checked for leaks by pressurizing with the gas and wetting with a
soap solution  or other commercially available  solutions.   Any  bubbles  forming on
the connections  indicate  a  leak.   Tighten or replace  any  leaking connections.   An
alternative leak check procedure for carrier gas is as follows:  Plug the outlet of
the analytical column.  Pressurize the tubing and connections.  Turn off the cylin-
der valve and  note  pressure on the regulator gauge and, if  equipped,  the GC pres-
sure gauge.  Any loss of pressure indicates a leak.   Locate  the leak using a soap
solution or  work backwards  through  the carrier  gas  flow path  disconnecting each
component and  plugging  the  carrier gas flow until  the  leak  is  located.   a similar
check should be made of the gas sampling valve,  sample loop,  and connections.
     It is advisable to take to the field an adequate  supply of spare parts, sep-
tums,  different  size  sample loops, extra  analytical  columns,  and  other related
equipment that may fail or deteriorate during the emission test.  The generation of
response factors for each  target  organic compound,  relative to  a  single organic
compound, should be  confirmed  in  the laboratory prior  to  going  to the field.  The
confirmation procedure  involving preparation  and  analysis  of calibration standards
containing multiple organic compounds is described in Section 5-1.6.

3.6  Reagents and Equipment

     The following  reagents  and equipment may be  required to conduct the emission
test depending on the sampling  method  selected.   These materials are generally ac-
quired from commercial  vendors.  Certification  of purity and/or analysis should be
obtained for adsorption  tubes,  calibration  and  zero gases, and liquid organic com-
pounds .

3.6.1  Charcoal  Adsorber - Check  to see  that the supply  of  charcoal  adsorbent is
sufficient to last for the entire field test period.

3.6.2  Adsorption Tubes - If adsorption tube  sampling is to  be conducted,  check to
see that the proper type of tube  has  been  obtained for collecting  the  target or-
ganic compounds.  Refer to  Table  B in the Methods  Highlights Section to determine
the proper adsorption material.   Check to see that the supply  of adsorption tubes
is sufficient  to conduct  the emission  test, including field  blanks and for desorp-
tion efficiency determinations.

3.6.4  GC Carrier Gas - Check  the GC operator's manual  and  the  GC column manufac-
turer to see that the GC  carrier gas type and grade are compatible with the GC and
the column.  Check  to see that  the supply of  carrier  gas is  sufficient to last the
entire field test period.

3.6.4  Auxiliary GC  Gases -  Check  to see  if the proper type  and grade of auxiliary
gases required by the GC detector have been obtained.   Consult with the GC detector
manufacturer to  determine the  proper  type  and  grade of auxiliary gases required.
Check to see that  the supply of auxiliary  gases  is sufficient  to  last  the entire
field test period.

3.6.5  Calibration Gases - Check to see if the correct calibration gases in the re-
quired range have  been obtained.    If  available,  commercial cylinder  gases  may be

-------
                                                                 Section No. 3-16.3
                                                                 Date June 30, 1988
                                                                 Page 31

used if their concentrations have been certified by direct analysis; cylinder gases
with tighter tolerances  on their concentrations are preferred.   Check  to see that
the supply of calibration gases is sufficient to last the entire field test period.

3.6.6  Calibration Gas Dilution System - Prior to each field trip or at the sign of
erratic behavior, any flow control valves or rotameters used in the dilution system
should be cleaned  according to the maintenance procedure  recommended by the manu-
facturer.   The rotameters  or  other metering devices  used with  a  single-stage or
two-stage dilution system  should be calibrated  prior  to  going to the field follow-
ing  the  procedures  described  in Subsection  2.2.   It is  advisable to  check the
dilution ratio  of  the dilution  system prior to  going to the  field following the
procedures described in Subsections 4.3.7 and 5*3-3-

3.6.7  Zero Gas  -  Check  to see that the zero gas meets  the  requirements for being
hydrocarbon-free (less than 0.1 ppmv of organic material as propane or carbon equi-
valent) .   Check to see that the supply of zero gas is sufficient to last the entire
field test period.

3-6.8  Audit Gases - Check to see that the required audit gases in the proper range
have been acquired.   Consult  Table  A in  the  Method Highlights  Section  for audit
gases available  from the EPA for the target organic compounds.    The availability
and ranges of audit gases can be determined by contacting:

        Environmental Protection Agency
        Environmental Monitoring Systems Laboratory
        Quality Assurance Division (MD-77B)
        Research Triangle Park, North Carolina 27711
        Attention:  Audit Cylinder Gas Coordinator

For audit gases  obtained from a commercial gas manufacturer,  check that the manu-
facturer has (1) certified the gas  in a manner similar  to the procedure described
in 40  CFR Part  61,  Appendix B, Method  106,  Section  5.2.3.1 and  (2)  obtained an
independent analysis of the audit cylinder that verifies that the audit gas concen-
tration is within 5% of the manufacturer's stated concentration.

3.6.9  Organic Compounds for Preparing Gaseous Standards - If gaseous standards are
to be prepared  in  the field,  check to see if the organic  compounds to  be used are
at least 99-9# pure  or,  if less  than  99-9#.  of known purity necessary to calculate
the gaseous standard concentration.   Record  the manufacturer's lot number for each
standard compound.

3-6.10   Equipment  for  Preparing  Gaseous Standards  by  Liquid or  Gas Injection-
Confirm that the Tedlar bags to contain the gaseous standards have been leak check-
ed following the procedures described in Subsection 3-5-9-   Check  to  see that the
syringes selected are gas-tight, cover the range needed (1.0- to 10-microliters for
liquids and 0.5 ml for gases), and are accurate to within 1#.  Confirm that the dry
gas  meter and  temperature gauge  have  been  calibrated  following  the procedures
described in Subsection 2.0.  Clean the midget impinger assembly with detergent and
tap water,  and then  rinse  with  deionized distilled water.   Check  the system for
leaks as  follows:    Assemble  the appropriate  system  for preparing  standards (see
Figure 5-5  for gaseous  materials  or Figure  5.6 for  liquid materials).    Fit the
injection port with  a new septum.  Fill  the  Tedlar bag and  pressurize the system
to 5 to 10  cm  (2 to 4 in.) Hg.  Any  loss  of pressure  after  10 minutes  indicates a

-------
                                                                 Section No. 3.1^-3
                                                                 Date June 30, 1988
                                                                 Page 32

leak. • Reject or repair the system if a leak is indicated.

3-7  Packing Equipment for Shipment

     The packing techniques described in this section are not requirements, but are
suggestions based  on previous field experience.  The  type  of packaging for equip-
ment going to the  field depends  on the mode of transportation.  Typically, packing
equipment for  transport by  a common carrier  will  require the  greatest  degree of
effort  to  ensure  the  equipment  arrives on-site  in its original  condition.   When
possible, delicate equipment  should  be packed in the original shipping containers.
For convenience, label all  containers  with the contents for easy identification in
the field.  The  most common mode of packing will be in a van or trailer,  where the
equipment will remain during transport.   More  sophisticated  test firms have trai-
lers or  trucks dedicated  to the  type of sampling being conducted.  These units are
often designed to  allow the test equipment and instruments  to remain set up during
transport.  This approach minimizes the time and effort  required to set  up before
and breakdown after  a test.  A  dedicated  test  vehicle provides a working environ-
ment that greatly  enhances the quality of work that can be performed.

3.7-1  Probe - Pack  the probe in a rigid case protected by polyurethane foam, poly-
ethylene  bubble-pack,  or  other  suitable  packing  material.    Seal  the  inlet  and
outlet of the probe  with  tape or other suitable  material.   Protect any protruding
glass ends from  breakage  by  insertion  into rigid plastic pipe  lined  with foam or
other packing material.

3.7.2   Teflon Tubing,  Sample Lines,  and Vacuum Lines   -  All  tubing, sample lines,
and vacuum lines should be  coiled and secured with  tape.   Coils should be large
enough not to crimp  tubing or excessively strain the  heat  sheath.   Seal  all open-
ings with tape.

3-7-3  Quick Connects, Flow Control  Valves and other Connectors  - All connectors,
valves,  and other  small parts should be packed in small parts cabinets, trays with
divided compartments, or  storage chests with labeled  drawers  to provide  quick and
easy access to the desired part.

3-7-4   Barometer  -  The  field  barometer  should  be packed in a  rigid container,
securely mounted in  rigid foam.   The  barometer case should be  packed  in a larger
box designated to contain delicate or fragile equipment.

3.7.5  Thermometers and Thermocouple Readouts - Thermometers and thermocouple read-
outs should be packed in  the  original  carrying case,  if possible.  Glass  thermome-
ters should be packed  in  a rigid tube to  prevent breakage.   These items, in their
smaller packing,  should also  be  packed  in  a larger  box designated to contain deli-
cate or fragile equipment.

3.7.6  Method 4 Equipment -  Method 4 equipment should be packed following the pro-
cedures recommended in Section 3-3-3 of this Handbook.

3.7.7  S-type Pitot  Tube  and Differential Pressure Gauge - The S-type pitot tube,
when not mounted on  the sampling probe, should packed  in a rigid case and wrapped
with polyurethane foam, polyethylene bubble-pack, or other suitable type of packing
material.  Seal  all  openings  with tape or other  suitable material.   The  differen-
tial pressure gauge, if  not  part of  a meter  box,  should be  mounted in  a rigid

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 33

housing. The gauge  should be wrapped with polyurethane  foam,  polyethylene bubble-
pack, or other  suitable  material,  and packed in a larger  box  designated for deli-
cate and fragile equipment.

3.7.8  Glassware - All glassware should be packed in the original shipping contain-
ers, if available, and stored together in a larger rigid container marked "Fragile!
Glass."  Otherwise, wrap the glassware with polyurethane foam,  polyethylene bubble-
pack, or other  suitable  material,  and pack in a rigid  foam-lined container marked
"Fragile! Glass."

3.7.9  Tedlar Bags - Preferably, transport the Tedlar bags to the field in individ-
ual  rigid  containers  used for sampling.   If  this  is not  possible,  pack the bags,
individually, in  corrugated  cardboard boxes with the connectors  secured such that
they do not contact and puncture the bags.

3.7-10  Sampling Pumps - Sampling pumps, if not mounted in a rigid housing suitable
for transport,  should be packed in a rigid foam-lined container.

3.7.11  Dilution  Interface System - The dilution interface  system  should be built
into in a rigid container suitable for shipment.

3.7.12   Gas Chromatograph System  -  The gas  chromatograph  and  ancillary systems
should be packed  in the  original  shipping container  for transport.   Although it is
not recommended,  the  GC  can be transported with out additional packaging in a van
or trailer provided the GC is secured properly against movement and other equipment
is not packed in  a manner where  it could fall on the instrument.  For transport in
dedicated  test  vehicles,  the instruments  should be  mounted  in  shock absorbing
devices.   All  gas lines  and analytical columns should be capped  to  prevent con-
tamination and/or oxidation during shipment.

3-7^13  Gas Cylinders -  All  gas  cylinders should be  transported with their protec-
tive cylinder heads securely attached.  The cylinders should be secured horizontal-
ly so that they do not roll together or vertically in a specially designed cylinder
rack.  Be aware of and adhere to all Federal, State,  and local regulations involv-
ing the transport of compressed and flammable gases,  particularly through tunnels.

3.7.1^  Liquid Organic Compounds  - Liquid organic  compounds  should be shipped with
the container top sealed with electricians tape and stored in a sealed plastic bag.
Packed each container in  its  original shipping box,  if  available.  Otherwise, wrap
each  container  individually with  polyurethane foam, polyethylene  bubble-pack,  or
other suitable material and place in a box designated for chemicals.

3.7.15  Dry  Gas Meters -  Dry  gas  meters not housed in  a  rigid meter box suitable
for  transport should  be wrapped with  polyurethane foam,  polyethylene bubble-pack,
or  other suitable material,  and packed in a  larger box designated  for delicate or
fragile equipment.

-------
                                                                 Section No.- 3.16,3
                                                                 Date June 30, 1988
                                                                 Page 34

     Table  3.1.  ACTIVITY MATRIX FOR PRELIMINARY  SURVEY  SAMPLING  AND  ANALYSIS

Characteristic
Apparatus Check
Barometer
Wet bulb/dry bulb
thermometers
Method 4 equipment
S-type pitbt tube
and differential
pressure
Probe
Teflon tubing
Quick connects
Glass flasks
High- vacuum pump
Tedlar or alumi-
nized Mylar bags

Acceptance limits
Within 2.5 mm
(0.1 in.) Hg of
mercury-in-glass
barometer
Within 1°C (2°F) of
a mercury-in-glass
thermometer
See Section 3.3-3
of this Handbook
See Section 3.1.3
of this Handbook
1. Clean; glass
liner, stainless
steel, or Teflon
inert to organics
2. Heating properly
if equipped with
heating system •
3. Leak free
New and unused
New or clean
Clean
Vacuum of 75 mm
(3 in . ) Hg absolute
Leak free; no
loss of pressure
after 30 seconds

Frequency and method
of measurement
Before each field trip
As above
Same as Section 3-3-3
Same as Section 3- 1-3
Before each field trip
following the proced-
ures described in Sub-
section 3 • 5 • 1
As above
As above
As above
As above
As above
As above
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20

Action if
requirements
are not met
Repair or replace
Replace
Same as Section
3-3.3
Same as Section
3.1-3
Repeat cleaning
Repair or replace
As above
Obtain new tubing
Clean according
to manufacturer ' s
recommendation
Repeat cleaning
of flasks
Repair or replace
As above
(Continued)

-------
Table 3.1  (Continued)
                                                                 Section No.  3-16.3
                                                                 Date June 30,  1988
                                                                 Page 35

Characteristic
Apparatus Check
Rigid containers
Direct pump
sampling system
Needle valve and
rotameter
Adsorption Tube
Procedure
Adsorption tubes
Personnel sampling
pump
Extraction solvent
Teflon tubing
On-site Measure-
ments and Sampling
Wet bulb/dry bulb
measurement

Acceptance limits
Leak free; no
loss of pressure
after 30 seconds
Leak free; no
loss of pressure
after 30 seconds
Clean
Proper type of
adsorption material
Calibrated
Proper type of
extraction solvent
New and unused
1 . Wet bulb wick .
moistened
2. Wet bulb temper-
ature stabilized.
3. Record wet bulb
and dry bulb
temperature

Frequency and method
> of measurement
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20
As above
Prior to each trip or
at the sign of erratic
behavior
Before each field trip
As above
Prior to extraction
of tubes for analysis
Before each field trip
Prior to each
measurement
During measurement .
.Immediately after wet
bulb temperature
stabilizes

Action if
requirements
are not met
Repair or replace
As above
Clean following
manufacturer ' s
recommendations
Replace with
proper type
Repair or replace
Replace with
proper type
Obtain new tubing
Moisten
Allow to
stabilize
Repeat
measurement
(Continued)

-------
Table 3.1  (Continued)
                                                                 Section No. 3.1^-3
                                                                 Date June 30, 1988
                                                                 Page 36
Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
On-site Measure-
ments and Sampling
Evacuated glass
flask sampling
1. Flask evacuated
to pump capacity

2. Assemble;  no
leakage
                    3- System purged up
                    to flask inlet

                    4. Stopcocks closed
                    and taped; flask
                    labeled

                    5- Flue gas tempera-
                    ture and static
                    pressure determined
Prior to sample
collection

Before sample col-
lection, visually and
physically inspect
all connections

Immediately prior to
sampling

Immediately after
sampling
                      Immediately after
                      sampling
Evacuate flask
                                                                  Check for leaks;
                                                                  repair system;
                                                                  repeat test
Purge system up
to flask inlet

Close and tape
stopcock; label
flask

Determine flue
gas temperature
and static
pressure
Purged glass
flask sampling
1. Assemble; no
leakage
                    3. Entire system
                    purged for 2 minutes
                    3. Stopcocks closed
                    and taped; flask
                    labeled

                    4. Flue gas tempera-
                    ture and static
                    pressure determined
Before sample col-
lection, visually and
physically inspect
all connections

Immediately prior to
sampling
                      Immediately after
                      sampling
                      Immediately after
                      sampling
Check for leaks;
repair system;
repeat test
Purge entire
system for 2
minutes

Close and tape
stopcock; label
flask

Determine flue
gas temperature
and static
pressure
(Continued)

-------
                                                                 Section No. 3-16.3
                                                                 Date June 30, 1988
                                                                 Page 37
Table 3.1  (Continued)
Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
On-site Measure-
ments and Sampling
Flexible bag sam-
pling procedure
1. Assemble using
Figure 4.4; no
leakage
                    2. Flow rate set to
                    0.5 1pm; purge sy-
                    stem up to bag inlet

                    3. Flue gas tempera-
                    ture and static
                    pressure determined

                    4. Bag labeled and
                    protected from
                    sunlight
Before sample col-
lection, visually and
physically inspect
all connections

Immediately prior to
sampling
                      Immediately after
                      sampling
                      Immediately after
                      sampling
Check for leaks;
repair system;
repeat test
Set flow rate
Purge system up
to flask inlet

Determine flue gas
temperature and
static pressure

Label bag and
protect from
sunlight
Adsorption tube
sampling procedure
1. Assemble using
Figure 4.9; no
leakage
                    2. Tubes capped,
                    labeled and stored

                    3. Flue gas tempera-
                    ture and static
                    pressure determined
Before sample col-
lection, visually and
physically inspect
all connections

Immediately after to
sampling

Immediately after
sampling
Check for leaks;
repair system;
repeat test
                                              Cap, label and
                                              store tubes

                                              Determine flue
                                              gas temperature
                                              and static press,
Preliminary Survey
Sample Analysis

Calibration
standards
(Continued)
1. Minimum of three
standards prepared
for each analyte

2. Sufficient peak
resolution achieved
(valley height <25%
of the sum of the 2
peak heights)
Prior to sample
analysis
                                          During multiple
                                          component standard
                                          analysis
Prepare three
standards for
each analyte

Vary GC operating
conditions and/or
change column
type

-------
                                                                 Section No. 3.16.,3
                                                                 Date June 30, 1988
                                                                 Page 38
Table 3.1  (Continued)
Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Preliminary Survey
Sample Analysis

Calibration
standards
3- Response for
consecutive repli-
cate injections of
each standard agree
within 5# of their
average response

4. Calibration curve
generated
                    5- Audit sample
                    (optional) analysis
                    results within 10%
                    of true value
During calibration
standard analysis
Repeat injections
                                          After calibration
                                          standard analysis
                      As above
                        Perform regres-
                        sion analysis and
                        plot curve

                        Repeat audit;
                        remake and
                        reanalyze
                        standards
Glass flask sample
analysis
1. Condensation in
sample flask
                    2. Flask not
                    pressurized

                    3. Condensation in
                    pressurized flask
                    after 10 minute
                    equilibration
                    4 .  Adequate resolu-
                    tion between peaks
                    achieved for peaks
                        of total area
Before sample analysis
                      As above
                      As above
(Continued)
                    5. Retention times
                    of consecutive in-
                    jections determined
                    and agree within 0.5
                    seconds or 1%
                      During sample analysis
                      After sample analysis
Heat flask to
flue gas or duct
temperature

Pressurize flask
                        Heat flask to
                        vaporize conden-
                        sate; if flask
                        already heated,
                        release pressure
                        and repressurize

                        Vary GC operating
                        conditions and/or
                        change column
                        type

                        Repeat analysis

-------
Table 3.1  (Continued)
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 39
Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Preliminary Survey
Sample Analysis

Flexible bag
samples
1.  Response for
consecutive repli-
cate injections of
each sample agree
within 5% of their
average response

2.  Stability of bag
samples acceptable
(second analysis
conducted an equal
number of days cor-
responding to the
the time elapsed
between sample col-
lection and first
analysis within 10%)
During sample analysis
Repeat analysis;
diagnose GC
problem
                                          After second analysis
                        Consider one
                        of the alternate
                        sampling methods
Adsorption tube
samples
1. Samples desorbed
for period specified
in referenced method
                    2. Response for
                    consecutive repli-
                    cate injections of
                    each sample agree
                    within 5% of their
                    average response

                    3- Desorption effi-
                    ciency >50%
Before sample analysis
                      During sample analysis
                      After sample analysis
Check referenced
method; desorb
for specified
period

Repeat analysis;
diagnose GC
problem
                        Evaluate more
                        vigorous desorp-
                        tion techniques;
                        Consider one of
                        the alternative
                        sampling methods

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 19"88
                                                                 Page 40
              Table 3.2.  ACTIVITY MATRIX FOR PRESAMPLING PREPARATION

Characteristic
Apparatus Check
Barometer
Wet bulb/dry bulb
thermometers
Method 4 equipment
S-type pitot tube
and differential
pressure
Probe
Teflon tubing
Quick connects
Sampling pump
Tedlar bags
Rigid containers

Acceptance limits
Within 2.5 nun
(0.1 in.) Hg of
mercury-in-glass
barometer
Within 1°C (2°F) of
a mercury-in-glass
thermometer
See Section 3. 3- 3
of this Handbook
See Section 3.1.3
of this Handbook
1. Clean; glass
liner, stainless
steel, or Teflon
inert to organics
2. Heating properly
if equipped with
heating system
3. Leak free
New and unused
New or clean
Leak free; adequate
delivery (> 1 Lpm)
Leak free; no
loss of pressure
after 10 minutes
Leak free; no
loss of pressure
after 30 seconds

Frequency and method
of measurement
Before each field trip
As above
Same as Section 3-3-3
Same as Section 3- 1-3
Prior to each trip
follow the cleaning
procedure described
in Subsection 3.5.1
Prior to each trip
As above
As above
As above
Prior to each trip
check with a rotameter
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20

Action if
requirements
are not met
Repair or replace
Replace
Same as Section
3.3.3
Same as Section
3-1.3
Repeat cleaning
Repair or replace
As above
Obtain new tubing
Clean according
to manufacturer's
recommendation
Repair or replace
As above
As above
(Continued)

-------
                                                                 Section No. 3-16.3
                                                                 Date June 30, 1988
                                                                 Page 41
Table 3.2  (Continued)
Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Apparatus Check

Direct pump
sampling system
Leakfree; no
loss of pressure
after 30 seconds
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20
As above
Needle valve and
rotameter
Clean
Prior to each trip or
at the sign of erratic
behavior
Clean following
manufacturer's
recommendations
Explosion risk
area sampling
system
Leakfree (no vacuum
loss after 1 minute)
Prior to each trip
Repair or replace
Heated bag
sampling container
1. Leakfree; no
loss of pressure
after 30 seconds
                    2. Heating properly
Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20
                      As above
As above
                        As above
Direct interface
sampling system
1. Heating properly

2. Leakfree (no
flow at rotameter
with probe plugged)
As above

As above
As above

As above
Dilution interface
sampling system
1. Heating properly

2. Flowmeters cali-
brated
                    3- Leakfree; no
                    loss of pressure
                    after 30 seconds
As above

Calibrate prior to
each test against a
bubble meter or
spirometer

Prior to each test
pressurize to 5 to 10
cm (2 to 4 in.) H20
As above

Calibrate
Gas chromatograph
equipment
Leakfree, opera-
tional, and suffi-
cient spare parts
for the duration of
the field test
Prior to field test
check system for
leaks,  access opera-
tional  condition, and
inventory spare parts
Consult the
operator's
manual
(Continued)

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1^88
                                                                 Page 42
Table 3.2  (Continued)
Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Reagents and
Equipment

Charcoal adsorber
Sufficient supply
Check supply prior to
each field test
Procure more
adsorbent
Adsorption tubes
Proper adsorbent,
tube size, and
quantity for test
Prior to field test
refer to Method
Highlights Section and
preliminary survey
results
Procure proper
adsorbent, tube
size, and
quantity
Gas chromatograph
carrier gas
Carrier gas compat-
ible to GC and
column; sufficient
quantity for test
Prior to field test
refer to operator's
manual or consult
with manufacturer
Procure compat-
ible carrier gas
in sufficient
quantity
Auxiliary GC gases
Proper type and
grade for GC detec-
tor; sufficient
quantity for test
Prior to field test
refer to operator's
manual or consult
with manufacturer
Procure proper
type and grade
of gases in suf-
ficient quantity
Calibration gases
Proper component(s)
and range; suffi-
cient quantity for
any on-site calibra-
tions
Prior to field test
refer to the prelim-
inary survey results
Obtain gases with
the proper compo-
nents in the
necessary range
and quantity
Calibration gas
dilution system
1. Rotameters clean
and calibrated
                    2. Dilution ratio
                    known (optional)
Prior to field test
examine and calibrate
following procedures
in Subsection 2.2

Prior to field test
establish the ratio
following the proced-
ures in Section 5-0
Clean and
calibrate
                                              Check dilution
                                              ratio prior to
                                              analysis
                                              (required)
Zero gas
Hydrocarbon-free
(<0.1 ppmv as pro-
pane or carbon equi-
valent) ;  sufficient
supply for test
Analyze or consult
manufacturer
Procure hydro-
carbon-free gas
in sufficient
quantity for test
(Continued)

-------
                                                                 Section No.  3.16.3
                                                                 Date June 30,  1988
                                                                 Page 43
Table 3.2  (Continued)

Characteristic
Audit gases
Organic compounds
for preparing
gaseous standards
Equipment for
preparing gaseous
standards
Packing Equip-
ment for Shipment
Probe
Teflon tubing,
sampling lines,
and vacuum lines
Quick connects ,
flow control
valves, and other
connectors
Barometer
Thermometers and
thermocouple read-
outs
Method 4 equipment
S-type pitot tube
and differential
pressure gauge

Acceptance limits
Required audit gases
in proper range
Target compound(s)
99.9$ pure or of
known purity
See Subsection
3.6.10
Protect with suit-
able packing
material
Coiled and taped;
openings taped
Stored organized
in containers
Packed in rigid foam
in a rigid container
Packed in original
container, if pos-
sible, or rigid
container
See Section 3-3.3 of
this Handbook
See Section 3.1.3 of
this Handbook

Frequency and method
of measurement
Prior to field test
contact EPA or vendor
(see Subsection 3-6.8)
Prior to field test
contact manufacturer
or vendor
See Subsection 3-6.10
Prior to each shipment
As above
As above
As above
As above
As above
As above

Action if
requirements
are not met
Acquire required
audit gas(es)
Procure 99-9$
pure compound(s)
or compound(s) of
known purity
See Subsection
3.6.10
Repack
Coil and tape
Repack
As above
As above
See Section 3.3.3
of this Handbook
See Section 3.1.3
of this Handbook
(Continued)

-------
                                                                 Section No. 3.16.3
                                                                 Date June 30, 1988
                                                                 Page 44
Table 3.2  (Continued)
Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Packaging Equip-
ment for Shipment

Glassware
Packed in original
shipping containers,
if available, or
suitable packing
material and marked
"Fragile"
Prior to each shipment
Repack
Tedlar bags
Packed in rigid sam-
pling containers, if
possible, or packed
individually in cor-
rugated boxes with
connectors secured
As above
As above
Sampling pumps
and dry gas meters
Mounted in a rigid
housing or packed in
rigid foam-lined
containers
As above
As above
Dilution interface
system
Built into a rigid
container suitable
for shipment
As above
Rebuild into
rigid container
or pack in suit-
able material
Gas chromatograph
system
Packed in original
shipping container,
secured properly in
van or trailer, or
mounted in a desig-
nated test vehicle
As above
Repack
Gas cylinders
Protective heads on,
secured in van or
trailer; transported
in compliance with
Federal, state, and
local regulations
As above
Repack;  check
Federal, state,
and local regu-
lations concern-
ing transport of
compressed gases
Liquid organic
compounds
Top sealed and pack-
ed in original ship-
ping container
 As above
Tape and repack

-------
                                                                Section No. 3.16.4
                                                                Date June 30, 1988
                                                                Page 1
4.0   ON-SITE MEASUREMENTS
    On-site  activities  include  transporting  the  equipment  to  the  test  site,
 unpacking and assembling the sampling and/or analytical equipment,  then conducting
 the sampling and/or analysis for the predetermined organic compound(s).  The qual-
 ity assurance activities for the  on-site measurements are summarized in Table 4.1
 at the  end  of this  section.   Copies  of all  field  data forms  mentioned  in this
 section are  in Subsection 3-16.12.   The  on-site measurements  checklist,  Figure 4.10
 at the end  of this section, provides the tester  with a quick method for  checking
 requirements during sampling.

 4.1  Transportation of Equipment  to the  Sampling Site

    The most  efficient means of transporting the equipment from ground level to the
 sampling site (often  above  ground  level)  should  be decided during  the  preliminary
 survey or by  prior  correspondence.   Care should  be taken to prevent damage to the
 equipment or injury to test personnel during the moving.   A clean "laboratory" type
 area free of excessive dust and organic compounds should  be  located and designated
 for preparing the sampling systems and conducting sample  recovery and analysis,  if
 applicable.

 4.2  Preliminary  Measurements and Setup

    Method 18 strongly  recommends  that  a  preliminary  survey  and/or  laboratory
 evaluation be  conducted prior to  sampling and  analysis.   Unless  adequate prior
 knowledge of  the  source  or information is  available,   the  presurvey procedures
 described in Subsection 3-0 on presampling operations should be  followed to select
 an acceptable sampling and  analytical approach.
    The accuracy of the sampling system(s) following handling and transportation  to
 the sampling site is  determined using a cylinder  gas  audit.   The integrity of the
 system(s) is  confirmed  after  setup by  conducting   the  individual  system  check
 described below for the applicable sampling method.   Preliminary measurements will
 always include determining  the stack dimensions and  the flue gas moisture.  Other
 measurements which may  be  made depending upon the requirements  of  the applicable
 regulations  and the source  operations include  a  flow rate determination,  velocity
 check, and stack  gas temperature  range measurement.
    One of the primary concerns for  any organic  sampling program must  be safety.
 The tester  should  always question  the  facility  representative  concerning general
 plant  safety requirements  and  safety  in   regard  to  sampling  at  the   selected
 sampling site.   Every  sampling  and analysis  protocol should address  the safety
 considerations involved  in performing  the  protocol.  Because there  are  numerous
 safety considerations involved  in organic sampling,  it is beyond  the scope of this
 Handbook to  discuss each one in detail.   However,  it cannot be  over-emphasised that
 the tester must always be aware of  the safety hazards.

 4.3  Sampling

    The following subsections discuss the procedures  for each Method 18  sampling
 technique.   At this point,  the tester has  selected  the proper  sampling technique
 and checked  the selected sampling  system.   If  this has not been  accomplished, the
 user should  refer to Subsection 3.0 prior to conducting the field test.

-------
                                                                Section No. 3.16.4
                                                                Date June 30, 1988
                                                                Page 2

   Because  of the  complexity in  sampling  organic compounds  from the  variety of
potential source  types,  only the more common problems  are  addressed  for each sam-
pling  method.   Recommended  quality  assurance/control  checks  and procedures  are
provided to assess  the suitability of the sampling technique for the samples to be
collected.  Because of  the relative compactness of the  equipment  and the low cost
of many of the sampling techniques, the tester may be able to utilize two different
sampling techniques at  the same  time with little additional effort.   The samples
from the  backup  or secondary  technique  are not analyzed if  the primary technique
proves satisfactory.  For example, the tester might easily  run  an adsorption tube
system as  a  backup to  an  evacuated  bag system.   At  some  facilities,  it  may be
necessary to  conduct two techniques  simply to  accurately measure  all  the organic
compounds of  interest.  The tester should always  be aware that a change in process
operations such as  raw materials, moisture content, operation mode, and temperature
can render a previously acceptable sampling technique unacceptable.
   The specific sampling system descriptions are provided below.

4.3.1   Evacuated Container Sampling  (Heated  and Unheated)  -  In  this  procedure,
sample bags are filled by evacuating the rigid air-tight containers that hold them.
The suitability of  the  bags for sampling should have  been  confirmed  by permeation
and retention checks using the  specific  organic  compounds of interest  during the
presurvey operations.   The  means  of transporting  the  bags to  the laboratory for
analysis within  the specified  time should also have  been determined.    Delays in
shipping and/or  analysis can  result  in  significant  changes in  concentration for
many compounds.
   On-site sampling includes the following steps:
     1. Conducting  preliminary measurements and setup.
     2. Preparation and setup of sampling system.
     3. Preparation of the probe.
     4'. Connection  of electrical service and leak check of sampling system.
     5. Insertion of probe into duct and sealing of port.
     6. Purging of  sampling system.
     7- Proportional sampling.
     8. Recording data.
     9- Recovering  sample and transportion to laboratory.

   Preliminary Measurements and  Setup - The  sampling  site  should be  checked to
ensure that  adequate electrical  service  is available.   The stack dimensions are
measured and recorded on a data sheet similar to the ones shown in Figures 4.1, 4.2
and 4.3.   The moisture  content of  the  flue gas  is  used to correct  the measured
concentrations to a dry basis.   It is typically measured prior to sampling using
wet bulb/dry  bulb thermometers  or  Method  4  (see Subsection 3-2); the determination
should be performed at a  time when process  operations  are like they will be during
final  sampling.   If the process  utilizes and  emits ambient  air,  a sling psychro-
meter may be used to measure the moisture content of the ambient air in the area of
process air uptake.  The moisture  content  value  is also used to  confirm that the
sampling approach selected is acceptable.
   Prior to final sampling,  the tester must determine if  the  final results are to
be presented  on  a concentration basis or a mass  emission basis.   If they will be
presented only on a concentration basis, only  the  concentrations  of  the specified
organics and the stack gas moisture content must be measured.  If the mass emission
rate of any  compound is  to be presented, the  flow rate of the stack gas  using a
velocity traverse must also be  determined.   In  this  case,  although not  required by
Method 18,  it is preferable that the sampling location be selected in accordance

-------
Plant Kubh&s C^IWm Co-
City Tt+sh'cr. *c.
Operator Ar}/)^> C^sse^
Date 3i-i -!»/&&
Run number' R'c.-l
Stack dia.
mm (in.) £ ;M.
Flowmeter calib. (Y)  Final leak check & m3/min (cfm)
Average (AP) 0.
Sample box number A//A
Pitot tube (Cr)
^).<&4-
Static press Q mm (in.) H20
Sampling
time,
min
0
£~
/O
'?
20
2fT
30
36T
to
4^
50
$-$
(pO
Total ifO
Clock
time,
24 h
M'AI-




10:4-1

Velocity head
mm (in. ) H20,
(AP)
0.
0.4(*
0.4b
OA(,
Avs 0>$?-
Initial flowmeter setting
Average stack temp (04
Barometric press
Flowmeter
setting
L/min (fti/minf
£>./S"2
o. A>?-
O.I&
0- &t>
0.1 (/O
O.l(*l
O-lbl-
0. /6£
&/Z2.
0. 141
0.l4-(*
0. 141,
0.141,
Avg 0,)5£
r(in.) HO Vacuum d
• /.02.
°* (0
7^,^ mm |ifrr)
stack
X" (°F)
^4
64-
64-
6»f
^f
^f
(?4
(?q-
W
bt
&4-
b*
(,4
Avg (,4,
probe
°C (°F)
»/A
)
(




Avg
uring leak check /£>

F) Sampling point locatio
Hg <&r>&r- o/^tf&tcjr
mm (in.) H20
n


Temperature
sample line
°C (°F)
A//A
\

(
\

/
Avg
readings
flowmeter box
°C (°F)
A//A
\
/
\
)
(

Avg
container
°C (°F)
A/4
\
/
(
)
/
(
Avg
Figure  4.1. . Field  sampling data form for  container  sampling.
T) O CO
(0 p (D
Oq rt O
(D (D ct

UJ t-i O
  c y

  § Z
     O
  UJ •
  O
   MD ON
   CO •
   00 -P^

-------
Plant Kubke-^ Cfa]L#\^ Co
City Tl^^i^ic flfC
Operator A~rjn* C &r\ter
Date / /^(, fbft
Run number^
Stack dia.
/yd i 	 w, /> a i -\

mm (in. ) (0 in.
Barometric
press ^f>T mm (iff.
Initial probe setting "°C
Sampling rate, £>./5~<
Sampling point locat
C£^?^y Of £>
7 Hg Sample loop volum
(°F) Sample loop temp
e /. O ml
nig "Q^(°V)
Z..,. L/min. (cfm) Column temperature:
ion
initial C,o/isf-»r)k /
/OO °C/min
{MC£~ . program rate fl//A . / °C/min
Dilution system: final — / /O0' °C/min
source flowrate £>. /5"2. L/min (cfm) Carrier gas flow
Meter box number A// A
Stack temp
/&D
2€"(°F)
Static press Q mm (in.) H20
Time of
injection
24 h
1011*
/04-7-
I/O]
H4\
/2-OI
'fl+O

Injection
number
^"z
fc'-l
*%?(,
*xee-l
ftl
£'L

Flow
source
ml/min
'*k
^
/«
/2
'f<\
IfZ.

^3 ml/min
diluent flowrate /OOl L/min (cfm) Dilution system check &jc '7.2-'%
diluent flowrate fiJ/A L/min (cfm) Final leak check
Dilution ratio 7-6" 2-5"

meter(s) s
diluent
ml/min
/oo?
/OOJ
loo?
/OOT-
1007-
1001-
/o&h
/DO*

0/<.
Vacuum @ check JO jpar^( in . ) H0 0

ettings
diluent
ml/min
\
1
")
(
)
(
I
stack
/eo
7^
/&(
ieo
/do
/e/
I&L

Temperat
probe
/80
/go
'%l
/e>4 •
tte
/do
I&5

ure readings
sample line
\&Q
IQo
/ez
'/£
/S
/?/
/^

injection port
2-/0
2-10
2-10
210
J-lt
JL/0
2^/3
•£JU no 0 Cfl
Oq rt O
(D CD rt
H-
-C- C-< O
C 3
(D Z
Figure 4.2.  Field sampling data form  for  direct interface sampling.
                                                                                   O
                                                                                   -  UJ
                                                                                   CO •
                                                                                   OO 4=-

-------
Plant
City
Operator
         TVjj/7'c  .
Run number	
Stack dia, mm  (in.)   (0 //?
Flowmeter calib.(Y)  /. / 2  J//m /i
Adsorption tube type:
 charcoal tube       X
 silica gel
 other
                                     Dilution system:  (dynamic)
                                      emission flowsetting    A//A
                                      diluent flowsetting
Meter box number
Pitot tube  (C)
                   PN-IZ.
Adsorp tion tube number
Average (aP)     Q.j-  mm  (in.) H20
                             A/2.
Static press   ()  mm  (in.) H20
Initial flowmeter setting
Average stack temp   7O
Barometric press  23. (/S mm  (in.) Hg
Dilution system: (static)
 emission flowsetting     A//A
Final leak check   (?  m3/min  (cfm)
Vacuum during leak check    /Q
                      mm  (in.) H20
                                                              C (°F)  Sampling point location
Sampling
time.
min
O
6T
/o
K
Z^
3D
3s"
40
45T
$&
fTS"
{,0
Total (fQ
Clock
time,
24 h
1222










132.2.

Velocity head
jjjar^in.) H20,
(AP)
04
04
04
0.4
O.f
0.4
04
04
04
04

04
Avg $ q.
Flowmeter
setting
L/min (&L/iafh)
/• 12-
A/2
//Z
I./2-
I.IZ.
1.17,
A/2
I.I-L
1.12-
/./£.
A/Z
/./I
Avg /, / 2.
Temperature readings
stack
2-e-(°F)
<>6
^6
W* 7
6^
7T^
"T,/"!
/ Cx
7/
^2-
7Z
^Z
1-2.
73
72-
Avg7/
probe, line
^og^op)
£g
6^
6f
^^
»
tfl
7-2
?2
?Z
7^x2
?3
-?^
Avg ^-/
adsorp . tube
-»€* (°F)
4ft
48 -
4^
4?
^&
5~-0
5c>
^0
57
5"/
r/
5"/
S'Z,
Avg 5-^1
meter
5*r(°F)
4e
*f-G>
4f
4^
I
50
ST?
5&
57?
5V
S~/
£"2-
Avg <^>

Vacuum
^paf. ( in . ) Hg
11
n
11
.11
n
If
IQ
11
11
n
n
let
Avg /^
                                                                                                             T3 O M
                                                                                                             CD (D n>
                                                                                                             oq rt o
                                                                                                             (0 (D ft
                                                                                                                 H-
                                                                                                             ui t-< o
                                                                                                               C 3
                                                                                                               to z
                                                                                                                 o
                                                                                                               uu >
                                                                                                               o
                   Figure 4.3-  Field sampling data  form for adsorption tube sampling.
                                                                                                               MD <
                                                                                                               00
                                                                                                               CO

-------
                                                                Section No. 3.16.4
                .  ,                                              Date June 30, 19§8
                                                                Page 6

with  Method 1.  If this  is  impractical,  it  should be  selected to  minimize flow
disturbances. The number and locations of sampling points for .the velocity traverse
are selected according  to Method  1  (see  Section 3-0.1  of this Handbook); the trav-
erse is conducted  according  to  Method  2  (see  Section 3-1 of this Handbook).  Note:
The Method  18 sampling will be conducted at a single point.
   Method  18  requires that  samples  be collected proportionally, meaning that the
sampling rate must be kept  proportional to the  stack  gas  velocity  at the sampling
point during  the sampling period.   If the process  has  a steady state  flow (con-
stant) , then the  flow rate does not  have to be varied during sampling.  The major-
ity of sources of organic emissions are of this  type because they use constant rate
fans. If the tester can confirm from the facility that-;'the source of interest has a.
steady state flow  (e.g., it uses a constant rate fan), then sampling can be conduc-
ted  at  a constant  rate  and  no concurrent  velocity measurements need  to be made.
If  it  is not  known whether  the  process is  steady state or  if it is  not steady
state, then velocity  measurements  (the velocity head)  must be made  at the point to
be  sampled.   This  can be  done   during the  preliminary  survey  or  before  final
sampling,  but  should be  done when  the  process operations  are like  they  will be
during the  final  sampling.   The average  velocity head  (pitot reading)  and range of
fluctuation  is  determined and then  utilized  to  establish  the  proper  flow rate
settings during  sampling.    If  it is found that the process is not  steady state,
then  the  velocity head  must be monitored  during  sampling to  maintain  a constant
proportion between the sample flow rate and the flow rate -in the duct.
   Select a total sampling  time greater  than or equal to the  minimum total samp-
ling  time  specified in  the  applicable emission standard.   The number  of minutes
between readings while sampling should be an integer.  It is desirable for the time
between readings to be such that the flow rate does not change more than 20% during
this period.
   If it was  determined from  the  literature  or the preliminary  survey laboratory
work that the sampling system must be heated during sample collection and analysis,
the average stack  temperature is  used  as. the  reference temperature  for the initial
heating of  the system and should be determined.  .Then,  the stack temperature at the
sampling point  is  measured  and  recorded during  sampling  to  adjust  the  heating
system just above  the stack  temperature  or  the dew point.   In addition,  the use of
a heated sampling  system  typically requires that the analysis be conducted on-site
since it is not  practical to maintain the sample  bag  at elevated temperatures for
long periods of time.                                         •

   Sampling  System Preparation  - Prepare  the  probe and sampling  train in  the
laboratory  area  (see  Figure  4.4).   First,  place a loosely  packed  filter of glass
wool in  the end  of the probe.  Attach a sample bag that  has  been  previously leak
checked to  the sample container lid.  Seal the inlet to the probe  and  the sample
container  lid  to  the  container body.   Transport the container  and probe  to the
sampling site.

   Proportional  Sampling -  Sampling  must be  conducted  at  a  rate  in  constant
proportion  to  the stack gas  flow  at  the sampling point. Thus, for  a  steady state
operation,   the sampling flow rate is  not  varied during the run. For  a non-steady
state process,  the  sampling flow rate  is varied in proportion  to  the changing
velocity.   The velocity  is monitored by measuring the velocity head  (AP) which is
linearly related  to  the square of the  velocity.  A recommended method  for deter-
mining proportional sampling rates is as follows:
   1. Conduct a single point velocity  check as previously  specified,  and determine
      the average velocity head" (APav ) to be sampled.

-------
                                                                           VENT
           STACK
            WALL
   FILTER
(GLASS WOOL)

       I
  REVERSE
  (3") TYPE
 PITOT TUBE
  TEFLON
SAMPLE LINE
                                       VACUUM LINE
            MALE QUICK
           CONNECTORS
NEEDLE
VALVE
                  FLOWMETE
                                                          PUMP
                                                                  CHARCOAL
                                                                    TUBE
                       NO CHECK
              PITOT MANOMETER
                           RIGID LEAKPROOF CONTAINER
                   Figure 4.4. Integrated bag sampling system.
                                                             *"0 U C/3
                                                             (0 (B (D
                                                             W ft O
                                                             (D (D rt
                                                                 H-
                                                             -^J «-i O
                                                               C 3
                                                               D
                                                               CD 2
                                                                 O
                                                               UJ .
                                                               O
                                                               -  OJ
                                                                                     VD CTi
                                                                                     00-
                                                                                     00 J^

-------
                                                                Section No.  3-l6v4
                                                                Date June 30,  1988
                                                                Page 8

   2. The average sampling flow rate for the test is  determined prior to the  start
      of the  run.   Typically,  the  average sampling  flow  rate is about  0.5  L/min
      which will yield approximately 30 liters of sample.  The flow rate chosen in
      the laboratory  should  fill the  bag  to about  three  fourths of  its capacity
      during the sample run.   The average  flow rate  chosen is  then assigned to the
      average velocity head measured.
   3. The flow rate to be  used  during sampling when  the velocity  head varies from
      the average is calculated  using the following  equation.
(—J
 »  a v cr *
                                         /2                             Equation 4-1
                                    a v g '
 where
      Q^ = Average sampling rate,  L/min (ft3/min),
      Qs = Calculated sampling rate,  L/min (ft3/min),
      AP = Actual velocity head,  mm (in.)  H20,  and
   APavg = Average velocity head,  mm  (in.) H20.

   4. Determine  the  rotameter setting for  the  sampling rate  (Qs )  from the  rota-
      meter calibration curve, and adjust  the rotameter accordingly.
Using this  procedure will  provide for the  correct sampling  rate  and the  proper
filling of  the sample  bag.   Follow  the  procedure below  to obtain an  integrated
sample.
   1. If  a  heating  system is  required,  turn  on  the  heating system  and set  at
      average stack temperature determined from the pretest measurements.
   2. Leak  check the sampling  train  just  prior  to  sampling by  connecting a  U-
      tube,   inclined  manometer,  or  equivalent  at  the probe  inlet  and pulling  a
      vacuum of  >_  10  in.  H20.   Close the  needle valve and  then turn the pump off.
      The vacuum should  remain  stable for  at  least  30  seconds.   If  a  leak  is
      found, repair  before proceeding; if  not, slowly  release  the vacuum  gauge.
      This leak check is optional.
   3. If  the  system  is  being heated, wait  for it  to come  to  the proper  tempe-
      rature.  Place  the probe in  the  stack  at  the sampling point:  centroid  of the
      stack or  no  closer  to  the  walls than  1  meter.    Seal 'the  sampling port  to
      prevent dilution of the stack gas by inleakage of ambient air.
   4. Disconnect the  flexible bag.   Purge  the  system by  turning on .the  pump  and
      drawing at  least  5  times  the  sampling system volume through the train,  or
      purge for 10 minutes, whichever is greater.
   5. Adjust the  flow rate to  the proper setting  based  on the velocity  pressure
      (during the purging, for non-steady  state  processes).
   6. Connect  the  flexible  bag  to   the  sampling  train  (the  connections  should
      ensure a leakfree system), and begin sampling.   The  rate must  remain propor-
      tional to the stack gas velocity for the total sampling time specified  by the
      standard of performance for the industry being sampled.
   7. Record all data required  (5 minute intervals, miniumum)  on the  field  samp-
      ling data  form (see  Figure  4.1).   The  flow rate  and sample train heating
      system should be  adjusted  after every pitot  and temperature reading to  the
      correct level.
   8. Disconnect and seal the flexible bag upon completion  of  sampling.  Take care
      not to dilute the contents  with ambient air.
   9. Label  each bag clearly and  uniquely  to  identify  it with its corresponding
      data form and/or run.  If the system is a heated system, the sample  bag must

-------
                                                                Section No.  3.16.4
                                                                Date June 30,  1988
                                                                Page 9

      be maintained at the stack temperature through sample analysis.

   Sample  Eecovery  and  Transport  to  Laboratory   -  Sample  recovery  should  be
performed so  as to  prevent contamination  of  the bag  sample  and  maintain  sample
integrity.  The bag  should remain leakfree, protected from  direct  sunlight,  main-
tained at a  temperature that  will  prevent condensation of  any of the  gases,  and
stored in a  safe  place to  prevent  damage or  tampering  prior  to analysis.   It is
recommended that bag samples  be  analyzed within two  hours of  sample collection,
however, many.of the organic compounds are  stable enough to allow a few days prior
to analysis.   Upon  completion of the  testing and  sample  recovery,   all  the  data
forms should be checked for completeness  and the  sample  bags reexamined for proper
identification.

   Common Problems  -  The  most  common  problems  encountered   with  bag  sampling
techniques are (1)  adsorption  of  the gases  on  the bag,  (2)  permeation of the gases
through the bag, (3)  reaction of gases in the bag,  (4)  condensation of the gases or
water vapor in the bag, and  (5) leaks  developing  in the  bag during testing,  trans-
port, and/or analysis.   As described previously in Subsection 3-0, the bags must be
checked for stability and retention of the  compound  in  the bag.   If the compound's
concentration significantly diminishes between the time the sample run is completed
and the time of analysis,  then the bag technique will have to be modified or rejec-
ted.   One modification  that can be used  to reduce  both  retention and/or condensa-
tion is addition of  a  heating system.   Heating is  generally applied  during sample
collection and  maintained  through  analysis.    However,  heating may  increase  the
permeation rate.   Another option is  the use of heat  lamps applied  to  the  sample
bags after  sample  collection  and during sample  analysis.   Two other  techniques
that have been used  to prevent condensation are  (1)  addition of a knockout trap to
remove  water  vapor  and heavy organics   from  the sample  stream,  and (2) use of
sorbents such as Tenax to remove  the high boiling point  organics.  The tester must
demonstrate  that  the  organic  compound(s)  of  interest  are  not   removed.
Alternatively, sample and/or water vapor  condensation may  be reduced  by the use of
the prefilled- bag  technique.    The prefilling  of  the bag  lowers the  concentration
of the organic and/or water vapor, thereby eliminating condensation.
   If the gases are  reacting in the bag,  then  the bag material  can be changed,  the
time between sample  collection and analysis reduced,  or  a different technique used
such as direct  interface  sampling.  Methods to reduce bag  leak  problems  are proper
construction of the  sample  bags,  conducting additional  runs, using a backup sample
collection technique such  as  an another bag sampling system or  an  adsorption tube
sampling system, and care  with handling  the sampling bags.   Also,  steel canisters
can be used in  place of the bags.  If the  organic  compounds are stable with time,
the use of steel canisters may better ensure the safety of the sample especially if
the samples must be air freighted to the laboratory for analysis.

4.3-2 Direct Pump  Sampling  - Direct pump  sampling is conducted  in a manner similar
to evacuated  container  sampling,  with  the exception that  the needle  valve and  the
pump are  located .between  the  probe and sample bag and  the sample exposed surfaces
of both must be constructed of stainless steel, Teflon or other material not affec-
ted  by  the  stack  gas  (see  Figure  4.5).   Due to  the additional  likelihood  that
sample may be lost in the needle  valve and pump,  it is recommended that the probe,
sample line,  needle valve,  and pump be heated.   If it has or can be shown that this
not a concern,  then  the heating  may  be  eliminated.   All  precautions, procedures,
data forms and  criteria  from  Subsection  4.3-1  above can be applied.   Ensure  that
the system has been adequately purged  before attaching the bag and sampling.

-------
                                                Stainless
                                               Needle Valve
                                                                                        Rotameter
    Filter
 (Glass Wool)
 Reverse
 (3") Type
Pilot Tube
                                                                 Teflon-Lined
                                                                  Diaphragm
                                                                     Pump
                       \
                                                                                      Protective Container
                                       Figure  4.5-   Direct  pump sampling  system.
*TJ C3 C/3
(D p] (D
Oq rt O
(D (D rr
     H-
M c_i o
O C 3

   (D 21
     O
   LO •
   O
   -  UO
                                                                                                                                      Oo
                                                                                                                                      CO

-------
                                                                Section No.  3.16.4
                                                                Date June 30, 1988
                                                                Page 11


4.3.3  Explosion Risk Area Bag  Sampling  -  Explosion risk area bag sampling is also
similar to evacuated container  sampling.   The  major difference is that no electri-
cal components can be used in the  explosion  risk -area.   As previously mentioned in
Subsection 1.0, the  first  option of the tester  is  to locate  the electrical equip-
ment (e.g., the pump) outside the  explosion  risk area and run a long flexible line
to the container.   If that option is not possible, an evacuated steel container may
be used as shown  in Figure 4.6.  This option  may involve a potential spark hazard
and must be checked  though the  plant  safety  officer.  No electrical heating of the
system will likely  be  allowed.   If an evacuated steel  container is used, the leak
check can be conducted outside  the explosion risk area  and the probe can be purged
with a hand squeeze pump.  The tester may wish to consider an alternative method of
sampling such as adsorption tubes and an intrinsically safe personnel sampling pump
or the syringe method.  The primary concern must be safety in an explosion risk area
and all operations must be outlined in writing and cleared through the Plant Safety
Officer.   The same, criteria as described above "for suitability of  the bag will
apply and must be met.

4.3.4  Prefilled Bag Sampling - The prefilled  bag sampling technique is similar to
the heated direct  pump sampling method.   The  major difference  is  that the sample
bag is prefilled with a known volume of nitrogen, hydrocarbon-free air, or cleaned,
dried ambient  air prior to  sampling  and the  volume of gas sampled  must be accu-
rately determined  (see Figure 4.5).  When using  a  flowmeter  or metering pump, the
maximum dilution that should be attempted is 10 tol.  Alternatively, a heated, gas
tight syringe may be used  to collect  the gas  at  the source and inject it into the
sample bag.  A heated, gas  tight syringe can be used for dilutions  of 5 to 1 when
the dilution  is  performed in the  syringe  and 50  to  1  when performed  in the bag.
The use of a heated, gas tight  syringe  should  follow the procedures shown below in
Subsection 4.3-5.    Both techniques  should  be  verified  in  the laboratory using
higher concentrations of calibration.... gases and must be within 10# of the calculated
value.   The technique  is  verified in the field  by  diluting-the audit gases in the
same manner as the stack gases  (see Subsection 8.0 for auditing procedures).
   Following are the recommended steps to conduct prefilled bag sampling:
   1.   The sampling  should be  conducted proportionally  as described  above  in Sub-
       section 4.3-1-  Calculation  of the  average sampling rate vs.  the average  P
       will be the  same  with the exception that  the volume of. the  prefilled inert
       gas must be taken into account.
   2.   The suitability of the  prefilled bag  sampling-  technique should  have been
       checked in  the laboratory.    This  would  include  calculating  the dilution
       factor required to obtain an acceptable sample concentration.
   3.   In the laboratory  area,  fill the sample  bag (previously leak checked) with
       the calculated  volume of inert  gas.    Because of the potential  for leaks,
       bags should  be  filled the, same  day ithey are used.   The inert  gas volume
       must be determined with  a calibrated  dry gas meter or  mass  flowmeter.  The
       bag should be sealed and taken to the sampling site.
   4.   At the sampling site,  the sampling system is leak checked without the
       sampling bag  attached.   Turn oh .the  heating system and  heat  the system to
       the stack temperature.   Connect a U-tube  H20 manometer or equivalent to the
       inlet of the probe.   After the system comes to the desired temperature, turn
       on the pump and pull  a vacuum  of about 10 in. of  H20.   Turn off the needle
       valve and  shut  off  the  pump.    If; there is no  noticeable  leak  within 30
       seconds,  then the system is leak free.
   5.   Place the probe  in the stack at the sampling point (centroid  or no less

-------
                                                                 PVC Tubing-
                                                                                            Directional
                                                                                              Needle
                                                                                              Valve
                                                                                                           Quick Disconnectors
Air Tight Steel Drum




      Sample Bag
                         Figure  4.6.  Explosion  risk area  sampling system  option  using an

                                       evacuated  steel container.
                                                                                                                      VD ON
                                                                                                                      00-
                                                                                                                      oo -c=-

-------
                                                                Section No.  3.16.4
                                                                Date June 30,  1988
                                                                Page 13

       than 1 meter from the wall) and seal  the  port  so there will be no inleakage
       of ambient  air.   Turn  on  the  pump  and  purge  the  system for  10  minutes.
       During the  time  that the  system  is purging,  determine and set  the  proper
       flow rate based on the Ap.
   6.   Turn off the pump and attach the sample bag.   Compare the heating system
   7.   The sampling will  be conducted proportionally.   The  stack  temperature and
       heating system  temperature should  be monitored and  recorded.    Record the
       data on the sampling data form (Figure 4.1).
   8.   At the conclusion of the run, turn off  the  pump and  remove  the probe from
       the duct.   Remove the bag and seal it.
   9.   Conduct a  final  leak check.  The  system  should pass  the  leak  check;  if it
       does not pass,  repeat the run.

4.3-5  Heated Syringe Sampling  -  The heated  syringe technique can be used with the
prior approval of the Administrator.   This technique should only be used when other
techniques are impractical.  The heated syringe technique requires on-site analysis
with three syringes collected and  analyzed for each run.   The requirements for the
use of the syringes are  the same  as  for  the  bag  with regard to the reaction of the
gases with time and the retention of the gases in the syringe.
   Following are the procedures recommended for the syringe sampling technique:
   1.  If heating  is  required,  then the  syringe must be encased in material that
       has a  high density  to maintain  the proper  temperature.   Alternatively, an
       external heating  system  can  be  used  that keeps the  syringe  at the proper
       temperature until  just  before use  and  to which the syringe  can be immedi-
       ately returned.
   2.  The access  port  should be  extremely  small to prevent inleakage of ambient
       air.   The  port  may be  covered with Teflon or other nonreactive material
       that will allow the  syringe to penetrate the material  for  sampling.
   3.  For  the  direct  injection  method  (no dilution),  place the  syringe  needle
       into the stack and  fill  and discharge the full  volume that will be sampled
       three  times.   Then, draw the emission sample  into the syringe,  immediately
       seal  the  syringe  and return  to   the  heating system,  if applicable.   The
       second and  third  syringes  are sampled at  equal  time intervals spanning the
       required  sample  (run) time.    The syringe  samples  must  not be  taken one
       immediately after another.
   4.  For  the  diluted  syringe  method, .the inert  gas  is  introduced into the
       syringe three  times and discharged.   Following this,  the proper volume of
       inert gas  is pulled into the syringe.  The  syringe  is then placed into the
       duct and  the  proper volume of  stack  gas  is added.   Immediately remove the
       syringe needle  from the  duct, seal the  syringe, and  return  to the heating
       system, if applicable.
   5.  For the  bag diluted  syringe  method,  the  bag  should be prefilled  with the
       proper volume  of  inert  gas.  The  sampling is conducted  as  described above
       and the sample injected into the bag through a septum.
   6.  Record the data on a field sampling data form (can adapt Figure 4.1).
   7.  Since  the  method   requires   a  proportional  sample   to be  collected,  the
       velocity head  (AP)   should be  recorded  at  about  the  same  time  that each
       sample  is  collected.    The  concentrations  can   then   be  mathematically
       corrected  to  provide an integrated value.   If the  process is  a constant
       source operation  (less  than  10% change in flow over  the  sampling period),
       it is not necessary to correct the measured values.

-------
                                                                Section No. 3.16..4
                                                                Date June 30, 1988
                                                                Page 14

4.3.6   Direct  Interface Sampling  - The  direct interface  procedure can  be used
provided  that  the moisture  content  of the  stack  gas does  not  interfere with the
analysis  procedure,  the physical requirements  of  the equipment can be  met at the
site,  and the source gas concentration is low enough  that  detector saturation is
not a  problem.   Adhere  to  all safety requirements  when using this method.  Because
of the amount of time  the  GC takes  to resolve the organic compounds prior  to their
analysis,  the GC  can  only  typically  make  three  analyses  in a  one-hour period.
Therefore,  the number of injections  in the direct  interface method is greatly lim-
ited  by  the  resolution time.   At  least  three injections  must  be  conducted per
sample run.
   Following  are  the   procedures  recommended  for  extracting  a  sample  from  the
stack,  transporting  the sample through a heated sample line, and introducing it to
the heated sample loop and the GC.   The analysis  of the  sample  is  described in
Subsection  5«0.
   1.   Assemble  the  system as  shown Figure 4.7, making all connections tight.
   2.   Turn on the sampling  system heaters.   Set the heaters to maintain the stack
        temperature as indicated by  the stack thermocouple.   If this temperature is
        above  the safe  operating temperature of the  Teflon  components,  adjust the
        heating system to maintain a temperature adequate to  prevent condensation of
        water and organic compounds.
   3-   Turn on  the  sampling pumps  and  set  the flow rate  at  the proper  setting.
        Typically 1 L/min is  used.
   4.   After  the system reaches the same temperature  as  the stack,  connect a U-
        tube H20  manometer  or  eqivalent  to  the inlet of the probe.   Pull a vacuum
        of  about  10  in.  of  H20, and shut  off  the  needle valve and  then the pump.
        The  vacuum should  remain stable for  30 seconds.   If  the  system leaks,
        repair and then  recheck the system.
   5.   Calibrate  the   system  as  described  in Subsection  5-0.   Repeat  until
        duplicate analyses are  within 5% of their mean value  (Subsection 5-0).
   6.   Conduct  the analyses of the  two audit  samples as described  in Subsection
        8.1.   The results must  agree  within  10% of  the true value (or greater,  if
        specified on  the cylinder).   If the results  do not agree,  repair the system
        and  repeat the analyses  until agreement is met or until approval is given by
        the  representative of the Administrator.
   7.    After  the audit  has  been successfully  completed,  place  the inlet  of  the
        probe at  the  centroid  of the duct,  or at  a  point  no  closer  to  the walls
        than 1  meter, and draw stack gas into  the  probe,  heated  line,  and sample
        loop.  Purge  the system  for a least 10 minutes.
   8.    Record the field sampling data on a form such as Figure 4.2.
   9-    Conduct the analysis  of the  sample as described in Subsection  5-0.   Record
        the  data  on the  applicable  data form (Figure  5-1. Subsection  5-0).   Ensure
        that the  probe  and  sample  lines  are maintained  at  0°C  to 3°C  above  the
        stack temperature (or a  temperature which prevents condensation).
   10.  Conduct the posttest calibration as described in Subsection 5-0.

4.3.7   Dilution  Interface Sampling -  Source  samples  that  contain a high concentra-
tion  of  organic materials  may require dilution  prior  to  analysis  to  prevent
saturating  the  GC detector.   The  apparatus  required for  this  direct  interface
procedure  is  basically  the  same as  described  above,  except a dilution  system  is
added between the heated sample line  and  the gas sampling  valve.   The apparatus is
arranged so that either  a 10:1 or 100:1 dilution of  the source gas can be directed
to the  chromatograph.   The description  of  the  apparatus is  presented in Subsection

-------
                                                                   MANOMETER
GLASS   1/2-in.
WOOL  TUBING
                                  TC
                               READOUT
  TC READOUT
OR CONTROLLER
         STACK WALL
                                                                          NEEDLE
                                                                           VALVE
                           tMPERATURE
                           CONTROLLER
                                                        HEATED
                                                       EFLON I INF
                                                            HEATED GAS
                                                          SAMPLING VALVE
                                                               INGC
            AUDIT
           SAMPLE
             IN
                                              CHARCOAL
                                              ADSORBER

                                          PUMP

                                          TO GC INSTRUMENT
FLOWMETER
                                      CARRIER IN
                                  Figure 4.7-  Direct interface  sampling system.
                                                                Tl O CO
                                                                03 CD ft)
                                                                OD, rr O
                                                                (D (D rt
                                                                   H-
                                                                h-1 <-4 O
                                                                Ul C D
                                                                 3
                                                                 (D Z
                                                                   O
                                                                 oo •
                                                                 O
                                                                 -  OJ
                                                                                                          OO
                                                                                                          00
                                                                                                            ON

-------
                                                                Section No. 3.16..4
                                                                Date June 30, 1988
                                                                Page 16

1.1.9  and the  pretest  calibration  of  the  apparatus  is presented  in Subsection
2.2.1.
   Following  are  the procedures  recommended  for extracting  a  sample  from  the
stack,  diluting the  gas to  the  proper level,  transporting the  sample  through a
heated  sample line,  and  introducing it to the heated sample loop  and the GC.  The
analysis of the sample is described in Subsection 5-0.
   1.  Assemble  the  apparatus  by connecting  the  heated box,  as  shown  in Figure
       4.8, between  the heated sample line  from  the  probe and  the  gas sampling
       valve  on the  chromatograph.    Vent the  source  gas  from  the  gas sampling
       valve directly to the charcoal filter, eliminating the pump and rotameter.
   2.  Measure  the stack temperature,  and adjust all  heating  units  to  a temper-
       ature 0°C to 3°C  above this temperature.  If the temperature is above the
       safe operating temperature of the Teflon components,  adjust  the heating to
       maintain  a temperature  high enough  to prevent  condensation of  water  and
       organic compounds.
   3.  After the heaters have come to  the  proper temperature, connect a U-tube H20
       manometer or eqivalent to the inlet of the probe.  Turn on  the pump and pull
       a vacuum  of about 10 in. of  H20.   Shut off the  needle  valve and then turn
       off the pump.  The vacuum reading should remain stable for  30 seconds.  If a
       leak is present,  repair  and then recheck the system.
   4.  Verify operation  of  the  dilution system by  introducing a calibration gas at
       the inlet of the  probe.  The diluted calibration gas  should be within 10% of
       the calculated value.    If  the  results for the  diluted  calibration gas are
       not within  10% of the expected values, determine  whether the GC and/or the
       dilution  system  is  in error.   If  the  analyses are  not  within acceptable
       limits because of  the  dilution  system,  correct  it  to  provide  the proper
       dilution  factors.   Make this correction by  diluting a  high concentration
       standard gas mixture to  adjust the dilution ratio as  required.
   5.  Verify the  GC  operation  using a low concentration standard by diverting the
       gas into  the  sample loop and bypassing the dilution  system as described in
       Subsection  5-1-   If  these analyses are not within acceptable limits, correct
       the GC by recalibration, etc.
   6.  Conduct  the analyses of the  two audit  samples  as described  in Subsection
       8.1 using  either  the dilution  system  or directly connect  the gas sampling
       valve as  required.   The results must agree within  10$ of the true value or
       greater value  if  specified on the  cylinder.   If  the results  do not agree,
       repair the  system and repeat the analyses  until agreement is  met or until
       approval is given by the representative of the Administrator.
   7-  After the  dilution  system  and GC operations  are  properly  verified and the
       audit successfully  completed,  place the  probe  at the centroid  of the duct
       or at a  point no  closer to the  walls  than 1 meter,  and  purge the sampling
       system for  at  least  10 minutes  at the  proper  flow rate.   Conduct the analy-
       sis of  the sample  as  described in  Subsection 5-0-   Record  the  field  and
       analytical  data on the applicable data  forms  (Figures k.2 and 5-1)-  Ensure
       that the probe, dilution system, and sample lines are maintained at 0°C to
       3°C above  the stack  temperature (or a  temperature  which  prevents conden-
       sation) .
   8.  Conduct the posttest calibration and verification of  the dilution system as
       described in Subsection 5-0.
   If the dilution system  is used for  bag sampling,  the procedures  for verifying
operation of  the dilution  system  will be the  same  as  shown above.   The diluted
calibration gas  will be collected in  a  bag  and  then verified.   Also  the  audit
samples will be collected in a bag and analyzed.   Acceptable results must be

-------
              Vent to Charcoal Adsorbers
i
Heated Line
from Probe X" \
Quick /" — ^\
Connect
Source
Gas Pump
1.5L/Min
Check Valve
Quick Connects
k i


Hoa

k 1
10:1
1

\
150 cc/Min
Pump
^— 3-Way
Valve:
) in 100:
' Positio

100:1
— Quick
Connects to
Gas Sample
Valve
-\ f
F
i

>,

3
150 cc/Min
Pump
• 1
X^

lorl Rnv at 19fl° P. nr "^niirr-o Tamnaraturo
I

1




.s— Flowmeters
.X"^ (On Outside
Jr of Box)
Flow Rate of
1350 cc/Min
TJ O 
CD pi (D
1 | "(B ft) rt
                                                                                      O
                                                                                      3
                                                           To Heated GC Sampling Valve
Figure 4.8.   Schematic  diagram of  the heated box  required for
             dilution  interface sampling.
OO
00

-------
                                                                Section No. 3-16..4
                                                                Date June 30, 1988
                                                                Page 18

obtained for the audit samples prior to analysis of the field samples.

4.3.8  Adsorption  Tube Sampling - Adsorption  tube sampling can be  used for those
organics specified  in  the Method Highlights Section,  Table B, and  for  other com-
pounds as  specified in  the  National Institute  of Occupational Safety  and Health
(NIOSH) methods.  The selection and use of adsorption tubes must be validated in
the laboratory  as  discussed  in Subsections 3-3  and 3-^ °r through  the  use of the
literature.  This check  will  include selecting the proper adsorption material, and
then checking the capacity, breakthrough volume, adsorption efficiency, and desorp-
tion efficiency.  The adsorption efficiency can be greatly affected by the presence
of water vapor  and  other organics in,  and  temperature of the stack gas.   If sam-
pling is conducted  for more  than one organic  compound,  the  adsorption and desorp-
tion efficiency checks must consider each.  Because changes  in process and control
equipment conditions can greatly  affect all of the parameters  stated  above,  it is
recommended as a standard operating procedure that more than one adsorption tube be
used.   The  first  tube  is analyzed as described in  Subsection  5-0.   If no problems
are found,  then the  second   tube  can be discarded.    If  problems with  the first
tube's adsorption efficiency are discovered, then the primary section of the second
tube  can  still be  analyzed  and  the results  included with  those of  the primary
portion of the first tube.
   Following are the recommended procedures for adsorption tube sampling:
   1.   The  sampling  system is  assembled  as shown  in  Figure 4.9-    The  adsorption
       tube(s) must  be maintained in a vertical  direction for sampling.   This is
       done to  prevent channeling of the  gases along  the side  of a  tube.   It is
       recommended  that the  sampling  probe  be eliminated  when  possible.    If  a
       sample probe  is used,  it  should be cleaned  prior  to its  initial  use with
       the  extraction  solvent.    Teflon tubing  should be used  for the  probe and
       sample line.
   2.   Just  prior  to  sampling, break  off  the ends  of  the  adsorption  tubes to
       provide  an  opening  at  least one-half  of  the internal  diameter.   Audit
       samples must  be collected on  the  adsorption tubes during  the  test program
       as  described  in Subsection 8.0.   Since  on-site analysis  is  typically not
       conducted when  using  adsorption tubes,  it  is  recommended  that two samples
       be collected  from each of  the two audit cylinders.   This  allows the tester
       a second chance to obtain the proper value for each audit cylinder.
   3.   Prior  to  sampling  and the collection  of  the  audit  samples,  the  sampling
       system  must  be  leak  checked  by  connecting  a  U-tube H20 manometer  or
       equivalent to the inlet  of the  sample  probe or adsorption tube.   Turn the
       pump on  and pull  a vacuum of about 10  in.  of H20.   Shut  off  the needle
       valve  and  then  turn off the  pump.   The vacuum  must remain stable  for 30
       seconds.  If a leak is present, repair and recheck the system.
   4.   If  the  flow  rate in the duct varies by  more than 10% during  the sampling
       period,  the  sample should  be collected  proportionally.    The  proportional
       sampling procedures will be the same as described for the bag sampling.  The
       only difference is that instead  of using  the volume  of the bag  as  the
       limiting factor to determine  the  average  sampling rate,  the  breakthrough
       volume is  the limiting  factor.    If the source is a  constant rate source
       (less than a  10%  change  in flow rate for the sampling  period), the samples
       can be collected at a constant rate.
   5.   Prepare the  field blank just  prior  to  sampling.   The  field blank  will be
       handled in be same  manner  as  the field samples  and should  be from  the same
       lot as the other adsorption tubes.
   6.   The flow rate meter must  have  been  calibrated in the laboratory prior to

-------
Supplemental

 Adsorption

   Tube

 as required)
                                                                    Probe
 Soap Bubble

  Flowmeter
(for calibration)
                                                                                                      T3 O C/3
                                                                                                      (a PJ (0
                                                                                                     OQ rt O
                                                                                                      fl> fl> ct
                                                                                                          H-
                                                                                                      H* CH O
                                                                                                     v^ C t3

                                                                                                        g ^
                                                                                                          O
                                                                                                        uo •
                                                                                                        O
         Figure  4.9.   Adsorption  tube  sampling  system.
                                                                                                        CO
                                                                                                        CO
                                                                                                           ON

-------
                                                             Section No. 3.16.4
                                                             Date June 30, 198&
                                                             Page 20

    the field trip  as  described in Subsection 2.1.  The  volume of sample coll-
    ected must be accurately known for adsorption tube sampling.
7.  The sample run  time  must be equal to or greater than that specified by the
    applicable regulation.   During each  sample  run,  the data should be recorded
    on the sample data form  (Figure 4.3 or equivalent).
8.  At  the  conclusion  of each  run,  conduct  another  leak  check  as  described
    above.   If  the system does not pass the  leak check, the run should be
    rejected, the leak located and repaired, and another run conducted.
9.  After completing  a successful leak  check,  remove the adsorption  tube from
    the holder  and  seal both  ends  with plastic caps.   The tubes  should  be
    packed lightly  with padding  to  minimize  the chance of  breakage.   If  the
    samples are to  be  held  for an extended period of  time,  they should be kept
    cool  to  reduce the  amount of migration  of the   organic  from  the primary
    section  to  the secondary  section.    Note:  Pack  the  tubes  separately from
    bulk samples to avoid possible contamination.
10.  It is recommended, that  at  the conclusion of  the  test,  the sample probe (if
    used)  be rinsed into a 20-ml glass scintillation vial with about 5 to 10 ml
    of the desorption  solvent.   This  sample will be analyzed  as  a check on the
    loss of  the organic  in  the probe  during sampling.   If more than 10$ of the
    total sample collected in  the  adsorption tubes is  present in the probe,  the
    samples should  be  rejected or  the sample catch adjusted  to account for the
    loss.   Alternatively, the  probe can  be  rinsed after  each  run and the rinse
    added to the desorption solvent prior to analysis.
11.  At  the  conclusion of  the test program,  check all  samples  to  ensure that
    they  are  uniquely  identified and check all  data sheets to  ensure that  all
    data has been recorded.

-------
                                                                Section No.  3.16.4
                                                                Date June 30,  1988
                                                                Page 21
WATER VAPOR CONTENT

Method 4

Reference Method conducted in proper manner 	 (Handbook Section 3-3i Method
   4, Figure 4.1)

Wet Bulb/Dry Bulb

Temperature readings taken when stabilized 	;	
   WB Temp °C (°F)	DB Temp °C (°F) 	  '
DIRECT OR DILUTION BAG SAMPLING

Apparatus

Pitot tube: Type S	   Other 	k, Properly attached
Pressure gauge: Manometer     '	  Other 	,  Sensitivity 	
Probe liner:  Borosilicate    '	 Stainless steel 	 Teflon 	._
   Clean  	,_, Probe heater (if applicable) on	  Glass wool filter
   (if applicable) in place  	.Stainless steel or Teflon unions used
   to connect to sample line 	
Sample line:  Teflon 	, Cleaned  	, Heated (if applicable)  	
     Tedlar 	  Other 	, Blank checked  	, Leak checked
   Reactivity check 	,  Retention check 	
Flowmeter: Proper range 	, Heated (if applicable) 	, Calibrated
Pump: Teflon coated diaphram 	,  Positive displacement pump 	,
   Evacuated canister 	,  Personnel pump	
Heated box with temperature control system: Maintained at proper temperature
Charcoal adsorption tube to adsorb organic vapors: Sufficent capacity 	
Dilution equipment: N2 gas 	, Hydrocarbon-free air 	    . Cleaned and
   dried ambient air 	, Dry gas meter 	
Barometer: Mercury 	, Aneroid 	, Other 	
Stack and ambient temperature: Thermometer 	, Thermocouple 	,
   Calibrated 	

Procedures

Recent calibration (if applicable): Pitot tube 	, Flowmeter 	,
   Positive displacement pump* 	, Dry gas meter* 	, Thermometer 	
   Thermocouple 	, Barometer 	
Sampling technique: Indirect bag 	, Direct bag 	, Explosion risk bag 	
   Dilution bag 	, Heated syringe 	, Adsorption tube 	.
   Proportional rate 	, Constant rate 	, Direct interface
   Dilution interface 	

*Most significant items/parameters to be checked.

                    Figure 4.10.  On-site measurements checklist.

-------
                                                                Section No. 3.16.4
                                                                Date June 30, 1988
                                                                Page 22
Figure 4.10 (Continued)
Filter end of probe (if applicable) and pitot tube placed at centroid of duct (or
  no closer than 1 meter to stack wall) and sample purged through the probe and
  sample lines* 	
Vacuum line attached to sample bag and system evacuated until the flowmeter
  indicates no flow (leakless)* 	
Heated box (if applicable) same temperature as duct* 	
Velocity pressure recorded and sample flow set 	
Proportional rate sampling maintained during run* 	
Stack temperature, barometric pressure, ambient temperature, velocity pressure
  at regular intervals, sampling flow rate at regular intervals, and initial and
  final sampling times recorded* 	
At conclusion of run,  pump shut off, sample line and vacuum line disconnected
  and valve on bag closed 	__
Heated box (if applicable) maintained at same temperature as duct until analysis
  conducted 	
No condensation visible in bag*    •-	
Sample bag and its container protected from the sunlight	
Audit gases collected in bags using sampling system*	
Explosive area bag sampling: (with following expections same as above) 	
Pump is replaced with an evacuated canister or sufficient additional line is added
  between the sample bag container and the pump to remove the pump from the
  explosive area 	
Audit gases collected in bags using sampling system* 	
Prefilled bag: Proportional rate 	 Constant rate 	
Dilution factor determined to prevent condensation*     '	
Proper amount of inert gas mete red. into bag through a properly calibrated dry gas
  meter*     '    _ •         .  .
Filter. end of probe (if applicable)  and pitot tube placed at centroid of duct (or
  no closer than 1 meter to stack wall) and sample purged through the. heated probe,
  heated sample line,  and heated flowmeter or positive displacement pump* _
Leak checked and- partially filled bag attached to sample line _
Stack temperature, barometric pressure, ambient temperature, velocity pressure at
  regular intervals, sampling rate at regular intervals, and initial and final sam-
  pling times recorded* _
Probe, sample line, and properly calibrated flowmeter or positive displacement pump
  maintained at the stack temperature* ___ _
Sampling conducted at the predetermined rate, proportionally or constant for entire
     *
  run
No condensation visible in probe, sample lines, or bag* _
At conclusion of run, pump shut off, sample line disconnected and valve on bag
  closed _
Sample bag and its container protected from sunlight _ _
Audit gases collected in bags using dilution system* _
Sample Recovery and Analysis

(As described in "Postsampling operations checklist," Figure 5-10)
*Most significant items/parameters to be checked.

-------
                                                                Section No.  3.16.4
                                                                Date June 30,  1988
                                                                Page 23

Figure 4.10 (Continued)

DIRECT AND DILUTION INTERFACE

Apparatus

Probe: Stainless steel 	,  Glass 	,  Teflon 	, Heated system (if
  applicable) 	, Checked     •	  •
Heated sample line: Checked* 	
Thermocouple readout devise for stack and sample line: Checked*
Heated gas sample valve: Checked*
Leakless Teflon-coated diaphram pump: Checked* 	'
Flowmeter: Suitable range 	
Charcoal adsorber to adsorb organic vapors 	___
Gas chromatograph and calibration standards (as shown in "Postsampling operations
  checklist," Figure 5.10)*^     » '

For dilution interface sampling only:
Dilution pump: Positive displacement pump or calibrated flowmeter with Teflon-
  coated diaphram pump checked* •	
Valves: Two three-way attached to dilution system-	
Flowmeters: Two to measure dilution gas, checked*	
Heated box: Capable of maintaining 120°C and contains three pumps, three-way
  valves, and connections, checked* 	
Diluent gas and regulators: N2 gas 	, Hydrocarbon-free air 	,  Cleaned air _,
  Checked 	

Procedures

All gas chromatograph procedures shown in "Postsampling operations checklist"
  (Figure 5.10)
Recent calibration: Thermocouples    •	, Flowmeter 	_, Dilution system
  (for dilution system only)* 	
Filter end of heated probe placed at centroid of duct (or no closer than 1 meter to
  stack wall), probe and sample line heat turned on and maintained at a temperature
  of 0°C to 3°C above  the source temperature while purging stack gas 	
Gas chromatograph calibrated while sample line purged* 	
After calibration, performance audit conducted and acceptable* '_	
Sample line attached to GC and sample analyzed after thorough flushing* 	
With probe removed from stack for 5 min, ambient air or cleaned air analysis is
  less than 5# of the emission results*	
Probe placed back in duct and duplicate analysis of next calibration conducted
  until acceptable agreement obtained* 	
All samples, calibration mixtures, and audits are analyzed at the same pressure
  through the sample loop* 	
Sample Analysis

(As shown in "Postsampling operations checklist," Figure 5.10)

*Most significant items/parameters to be checked.

-------
                                                                Section  No.  3.16.if
                                                                Date  June  30,  1988
                                                                Page  24
Figure 4.10  (Continued)
If a dilution system is used, check  the following:
With the sample probe, sample line,  and dilution box heating  systems  on,  probe  and
  source thermocouple inserted  into  stack and all heating systems  adjusted  to a
  temperature of 0°C to 3°C  above  the stack temperature 	
The dilution system's dilution  factor is verified with a high concentration gas of
  known concentration (within 10%) 	
The gas chromatograph operation verified by diverting a low concentration gas into
  sample loop
The same dilution setting used  throughout the run 	
The analysis criteria is the same  shown as for the direct interface and in  the
  "Postsampling operations checklist," Figure 5.10

ADSORPTION TUBES

Apparatus

Probe: Stainless steel 	,  Glass 	, Teflon 	, Heated
  system and filter (if applicable)  	
Silica gel tube or extra adsorption  tube used prior to adsorption  tube when
  moisture content is greater than 3 percent	
Leakiess sample pump calibrated with limiting (sonic) orifice or flowmeter  	
Rotameter to detect changes  in  flow 	
Adsorption tube: Charcoal (800/200 mg), Silica gel (1040/260 mg)     .	
Stopwatch to accurately measure sample time	•

Procedures

Recent calibration of pump and  flowmeter with bubble meter 	
Extreme care is taken to ensure that no sample is lost in the probe or sample line
  prior to the adsorption tube	
Pretest leak check is acceptable (no flow indicated oh meter)	
Total sample time,  sample flow rate,  barometric pressure,  and ambient temperature
  recorded 	
Total sample volume commensurate with expected concentration and recommended sample
  loading factors 	
Silica gel tube or extra adsorption tube used prior to adsorption  tube when
  moisture content is greater than 3 percent 	'   	
Posttest leak check and volume rate meter check is acceptable  (no  flow indicated on
  meter,  posttest calculated flow rate within 5 percent of pretest flow rate)

Sample Analysis

(As shown in the "Postsampling operations checklist," Figure 5.10)
*Most significant items/parameters to be checked.

-------
                                                                Section No. 3.16.4
                                                                Date June 30,  1988
                                                                Page 25
               Table  4.1.   ACTIVITY MATRIX FOR ON-SITE MEASUREMENTS
 Characteristic
  Acceptance limits
 Frequency and method
    of measurement
Action if
requirements
are not met
 Preliminary de-
  terminations
  and measure-
  ments
If final results on
concentration basis,
determine the. mois-
ture content of stack
gas

If final results on
emission rate basis,
determine moisture
content and flow
rate of stack gas
                   If process has >IQ%
                   variation in APa v e,
                   sampling must be con-
                   ducted proportionally
                   If preliminary survey
                   or information showed
                   a heating system nec-
                   essary for sampling,
                   determine stack gas
                   temperature,  Ts

                   Determine stack
                   dimensions

                   Select sampling time
                   >^ minimum total
                   sampling time in
                   applicable emission
                   standard; number or
                   minutes between
                   readings should be an
                   integer
 Once each field test;
 use wet bulb/dry bulb
 thermometer, Method 4,
 or sling psychrometer
 See above for moisture
 content; for flow
 rate, once each field
 test using Method 1
 location, if possible,
 and Method 2 proce-
 dures

 Determine before
 test by measuring
 APaye and range of
 fluctuation; if re-
 quired, use APave and
 AP measured during
 sampling to vary
 sampling flow rate to
 sample proportionally

• Prior to and during
 sampling
                       Prior to sampling,
                       using tape measure

                       Prior to sampling
 Complete
                                                                   Complete
                                                Complete or
                                                repeat sampling
                                                Complete or
                                                repeat sampling
                           Complete


                           Complete
(Continued)

-------
Table 4.1 (Continued)
                                                                Section No. 3.16.4
                                                                Date June 30, 198*
                                                                Page 26
 Characteristic
  Acceptance limits
    Frequency and method
       of measurement
                        Action if
                        requirements
                        are not met
 Evacuated con-
  tainer sampling
1. Assemble system
   using Fig. 4.4;
   leakage
no
                   2. Minimum vacuum
                      of 10 in.  of H20;
                      stable for 30 s
                   3. Heating system,  if
                      used,  between 0°
                      and 3°C above
                   4. Locate probe tip
                      at centroid of
                      stack or no closer
                      than 1 meter to
                      walls of stack

                   5. Purge probe and
                      sample system,
                      5 times system
                      volume or 10
                      minutes ,  which-
                      ever is greater

                   6. Sample propor-
                      ionally based on
                      AP0.,_ and moni-
                        3 V C
                      tored AP
Before sample collec-
tion, visually and
physically inspect
all connections
                       Before sample collec-
                       tion; use a H20-
                       filled U-tube
                       manometer or equiva-
                       lent
                       Confirm prior to and
                       monitor during sam-
                       pling using tempera-
                       ture sensor(s)

                       Prior to sampling;
                       determine using stack
                       dimensions
                       Before sample collec-
                       tion; with bag
                       unattached, turn on
                       pump
                       Throughout sampling
Check for leaks,
repair system;
repeat check
                            Check system
                            for leaks;
                            check pump,
                            joints, and
                            valves for source
                            of leak; repair
                            and recheck

                            Adjust heating
                            system
                            Reposition
                            Repeat purge
                            Repeat test
 Direct pump
  sampling
(Continued)
1. Assemble system
   using Fig.  4.5;
   sample exposed
   components  of
   Teflon,  stain-
   less steel,  etc;
   no leakage
    Before sample col-
    lection,  visually
    and physically in-
    spect all equipment
    and connections
                        Check for leaks,
                        repair system;
                        replace inappro-
                        priate components

-------
Table 4.1 (Continued)
                                                                Section No. 3.16.4
                                                                Date June 30, 1988
                                                                Page 27
 Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
 Direct pump
  sampling (cont)
2. Follow steps 2-6
   for evacuated bag
   sampling
As above
As above
 Explosion risk
  area bag
  sampling
1. Assemble system,
   Figure 4.6 is one
   option; no elec-
   trical compo-
   nents in explo-
   sion risk area;
   no leakage

2. Leak check as
   above outside
   explosion risk
   area

3. Purge probe with
   a hand squeeze
   pump changing
   volume at least
   5 times

4. Follow steps 4
   and 6 for evac-
   uated bag samp-
   ling

5. Clear all oper-
   ations in writ-
   ing through
   Plant Safety
   Officer
As above
As above
                                          As above
                        As above
                                          As above
                        As above
                                          As above
                        As above
                                          Prior to working
                                          in explosion risk
                                          area
                        Complete
  Prefilled bag
    sampling
(Continued)
1. Assemble system
   using Fig. 4.5;
   need calibrated
   flowmeter in-line

2. Calculate accept-
   able dilution
   factor

3- Leak check bag
As above for
evacuated bag
sampling
                                          Prior to sampling
                                          Prior to filling
As above for
evacuated bag
sampling
                        Complete
                        Repair or replace

-------
Table 4.1 (Continued)
                                                                Section No. 3.16.4
                                                                Date June 30, 19*88
                                                                Page 28
 Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
  Prefilled bag
   sampling (cont)
4. Fill bag with
   known volume of
   inert gas
                    5-  Leak check system
                       at stack temper-
                       ature ,  minimum
                       vacuum of 10 in.
                       of H20, stable
                       for 30 s

                    6.  Follow steps 3~6
                       for evacuated bag
                       sampling

                    7•  Determine volume
                       of gas sampled
                       accurately
Prior to sampling;
use calibrated dry gas
meter or mass flow-
meter

Before sample collec-
tion without bag
attached; use U-tube
H20-filled manometer
or equivalent
                      As above
                      During sample collec-
                      tion; use flowmeter
                      or metering pump (max.
                      dilution 10 to 1) or
                      heated syringe, (see
                      below (max. dilution
                      50 to 1)
Complete
                                              Locate leak,
                                              repair or
                                              replace compo-
                                              nents ,  and
                                              recheck
                        As above
                        Complete
  Heated syringe
   sampling -
   direct injec-
   tion
1.  Check syringes
   for compound re-
   tention and re-
   action

2.  Seal port to pre-
   vent inleakage
   of ambient air

3-  Place needle in
   stack at sample
   point, pull and
   discharge sample
   volume three
   times
See Subsection 1.0
                                          Visually check
                                          Prior to sampling
Complete
                        Reseal and re
                        check
                        Complete
 (Continued)

-------
Table 4.1 (Continued)
                                                                Section No. 3.16.4
                                                                Date June 30, 1988
                                                                Page 29
 Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
  Heated syringe
   sampling - dir-
   ect injection
   (cont)
   Seal after pull-
   ing sample vol-
   ume, return to
   heating system,
   if necessary;
   monitor heating
   system tempera-
   ture

   Choose sample
   volumes to sample
   proportionally;
   monitor AP, if
   necessary (>10#
   change in flow
   over sampling
   period)

   Take second and
   third syringe
   samples at equal
   time intervals
   spanning the
   required sampling
   time in applica-
   ble emission
   standard
For each sample
collection; use
temperature sensor
Complete
                                          During sample collec-
                                          tion;  use pitot tube
                        Repeat sampling
                                          During field test
                        Repeat sampling
  Heated syringe
   sampling -
   dilution
   method
  (Continued)
1. Follow same steps
   as for heated
   syringe - direct
   injection, except
   prefill bag (see
   steps 2-4 in pre-
   filled bag samp-
   ling) and inject
   gas in heated
   syringe through
   bag septum
As above
As above

-------
Table 4.1 (Continued)
                                                                Section No.  3.16.4
                                                                Date June 30,  1988
                                                                Page 30
 Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
  Direct interface
   sampling
1.  Assemble system
   using Fig.  4.7;
   no leakage
                    2.  Heating system
                       between 0°  and
                       3°C above T
                    3.  Set flow rate at
                       1 L/min

                    4.  Leak check system
                       at stack temper-
                       ature at minimum
                       vacuum of 10 in.
                       of H20; stable
                       for 30 s

                    5-  Calibrate system;
                       duplicate analy-
                       ses within 5# of
                       their mean

                    6.  Analyze audit
                       cylinders;
                       results within
                       10% of true value

                    7.  Follow steps 4
                       and 5 of evacu-
                       ated bag
                       sampling

                    8.  Analyze samples
                       and conduct
                       posttest
                       calibration
Before sample collec-
tion visually and
physically inspect
all connections

Confirm prior to and
monitor during
sampling using temper-
ature sensors

Prior to sampling
Check for leaks;
repair system;
repeat check
                                              Adjust
                      During sampling;  use
                      a U-tube H20  mano-
                      meter or equivalent
                      See Subsection 5.0
                      See Subsection 8.0
                      As  above
                      See Subsection 5.0
                                              Complete
                        Check system for
                        for leaks; repair
                        and recheck
                        Identify
                        problems; recal-
                        ibrate and check
                        Reject samples
                        and rerun test
                        As above
                        Complete
   (Continued)

-------
Table 4.1 (Continued)
                                                                Section No. 3.16.4
                                                                Date June 30, 1988
                                                                Page 31
 Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
  Dilution inter-
   face sampling
1. Follow all steps
   for direct inter-
   face sampling
   with addition
   of steps below

2. Assemble with
   dilution system
   in line, see
   Figure 4.8

3. If Teflon com-
   ponents cannot
   withstand stack
   temperature,
   heating system
   should be set to
   prevent conden-
   sation

4. Verify dilution
   system to within
   10% of expected
   value
As above
As above
                                          As above
                        As above
                                          Prior to and during
                                          sampling
                        Adjust
                                          Prior to sampling; use
                                          a calibration gas
                        Pinpoint problem
                        to dilution
                        system or GC;
                        repair and
                        recheck; adjust
                        dilution, if
                        necessary
  Adsorption tube
   sampling
1. Assemble system
   using Figure 4.9
                    2. Break off ends of
                       adsorption tubes;
                       maintain in ver-
                       tical position
                       for sampling

                    3. Follow step 4 for
                       direct interface
                       for leak check
Before sample coll-
ection, visually and
physically check all
connections

Just prior to samp-
ling; during sampling
                      As above
Check for leaks,
repair, and
recheck
                                              Complete and
                                              check
                        As above
(Continued)

-------
Table 4.1 (Continued)
                                                                Section No. 3.16.4
                                                                Date June 30, 1988
                                                                Page 32
 Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
  Adsorption tube
   sampling (cont)
4. Conduct sampling
   proportionally,
   if the flow rate
   varies more than
   10%

5- Determine samp-
   ling time, >^ re-
   quirement of
   applicable
   emission standard

6. Run field blanks
                    7. Perform posttest
                       leak check as
                       above
During sampling
Repeat test
                                          Prior to sampling
                        Complete
Once each set of
samples

As above
                                                                  Complete
                                              Reject sample,
                                              rerun test
 Sample
  recovery
(Continued)
1. If applicable,
   remove samples
   from sampling
   system

2. Protect bag samp-
   les from sunlight
   and maintain at a
   temperature which
   will prevent con-
   densation

3- Analyze bag sam-
   ples within two
   hours of sampling

4. For adsorption
   tube samples,
   perform at least
   one probe rinse
   with desorption
   solvent to con-
   firm that <10% of
   sample is col-
   lected in probe
Following sampling
                                          Following sampling
Complete
                        Complete
                                          Following sampling
                                          Following sampling;
                                          analyze sample with
                                          GC
                        Complete
                        Adjust sample
                        values to
                        account for probe
                        catch

-------
                                                                Section No. 3.16.4
                                                                Date June 30, 1988
                                                                Page 33
Table 4.1 (Continued)
 Characteristic
Acceptance Limits
Frequency and method
   of measurement
requirements
are not met
  Sample
   logistics
Properly label
all bags, contain-
ers, tubes, etc.
                     Record all data on
                     forms in Figs. 4.1,
                     4.2, and 4.3 and
                     5-1
Visually check
each sample
                      As above
Complete the
labeling
                         Complete the
                         data records

-------
                                                              Section No.  3-16.5
                                                              Date June 30,  1988
                                                              Page 1
5.0   POSTSAMPLING  OPERATIONS
       The postsampling operations for Method 18 include preparation of calibration
standards appropriate for the sampling  technique used,  determination of desorption
efficiency and  collection efficiency  for adsorption  tubes  (if used),  adsorption
tube sample preparation,  sample analysis, and determination of  acceptable  resolu-
tion and  precision.   See Subsection  3-1-5  for postsampling operations  related  to
velocity determinations and Subsection 3-3-5 for postsampling operations related  to
determination of  the  flue gas  or  duct  moisture.   Figure  5-10 at the end  of  this
section provides a checklist for monitoring the postsampling operations.   Table 5-1
at the  end  of the section  summarizes the quality assurance activities  associated
with the postsampling operations.

5.1  Preparation of Calibration Standards

       Calibration standards are to be  prepared prior  to sample analysis following
the procedures  described in the  following  subsections.   Refer  to Table E  in the
Method Highlights Section for recommendations on the procedures suitable for selec-
ted compounds.   Note  that there are two basic types of standards,  gaseous  or  liq-
uid;  the  type prepared  depends on the  type of sample collected.    Gaseous cali-
bration standards will  be needed prior to the analysis  of preliminary survey  sam-
ples collected  in glass  flasks  or bags, and final samples  collected in bags or  by
direct and dilution  interface  sampling.   There are  three  techniques for preparing
gaseous standards,  depending on availability  and  the  chemical  characteristics  of
the standard compound(s); gas cylinder  standards may also  be used directly,  if the
proper  concentration  ranges are available.   Liquid calibration standards  are re-
quired  for  the  analysis of adsorption  tube samples  from the  preliminary  survey
and/or the final sampling, as well as to determine the desorption efficiency; there
are two techniques for  preparing liquid calibration standards.  The concentrations
of  the calibration  standards   should bracket  the  expected concentrations  of the
target compound(s) at  the source being tested.  Specific  procedures for preparing
and analyzing each type of standard are described below.
       For each  target  compound,  a minimum of three different standard concentra-
tions are required to calibrate the GC.   An exception  to this  requirement involves
developing relative response factors  for  each  compound  to  be  tested as compared  to
a single  organic compound.   Once  in  the field,  the GC  is calibrated for all target
compounds using  the  single  organic.   The validity of this  procedure must be first
be proven in  the laboratory prior to the test.   To save  time,  multiple component
standards can be prepared and analyzed provided the elution order of the components
is known.
       It is  recommended that  the  linearity of the calibration curve  be  checked
comparing the actual  concentration of the calibration  standards  to  the concentra-
tion of the standards  calculated using the standard peak  areas  and  the linear re-
gression  equation.   The recommended  criteria  for  linearity is  for  the calculated
concentration for each standard be within J% of the actual concentration.
       After establishing the GC  calibration curve,  an  analysis of  the  audit  cyl-
inder is performed as described in Subsection 8.1.   For an instrument drift check,
a  second  analysis  of  the  calibration  standards  and  generation  of  a  second
calibration curve is  required  following sample analysis.  The area  values  for the
first and second analyses of each standard must be within  5% of their average.   If
this criterion cannot be met,  then the sample values obtained using the first and
second  calibration  curves  should  be  averaged.    In addition,  if  reporting  such

-------
                                                              Section No. 3.16.5
                                                              Date June 30, 1988*
                                                              Page 2

average  values  for the samples  is  warranted,  an additional  analysis  of the audit
.cylinder should  be performed.   The average of  the  audit values obtained using  the
two calibration  curves should be reported.

5.1.1   Commercial Gas Cylinder Mixtures - Commercial  gas  cylinder mixtures can be
used  provided that the  cylinders  have . been certified by direct  analysis  and  the
proper  concentrations for  the  emission test  can be  obtained.    Calibrate  the GC
using gas cylinders by the  following procedure:
     1.  Secure  the three cylinders near  the  GC and remove  their protective caps.
         Attach an appropriate regulator that  is equipped with a flow control valve
         to the lowest concentration standard.
     2.  For preliminary survey  sample  analysis,  establish the proper GC conditions
         based  on  the  referenced conditions in Table D in  the  Method Highlights
         Section,  previous experience,  or  possibly,  if  the plant being tested has a
         laboratory, the laboratory personnel.  For final sample analysis, establish
         the optimum GC conditions  determined during the preliminary survey sample
         analysis.
     3.  Attach a quick connect  or equivalent,  compatible  to  the  connection on  the
         Tedlar bag or the  interface sample line, to the gas  sampling valve on  the
         GC.
     4.  Connect  a length of Teflon  tubing  to  the flow  control valve on the regula-
         tor and  connect the other  end, using  a compatible connector,  to  the gas
         sample valve.
     ,5.  With the  gas sampling valve in the load position and the flow control valve
         open,  open the valve  on the cylinder and adjust the  pressure regulator to
         deliver  a flow of lOOcc/min  through  the sample loop,  determined by a rota-
         meter or  other flow sensing device on the loop outlet.
     6.  Allow the sample loop to be  flushed  for 30  seconds,  then turn off the flow
         control valve.
     7.  Allow the sample  loop  to return  to  the same pressure  that  will be exper-
         ienced during sample  analysis, determined  with a. manometer  or equivalent
         connected  to  a tee  on  the  outlet  of the loop,  and  immediately switch the
         valve to  the inject position.
     8.  Note  the  time of  the  injection on the  strip chart recorder and/or actuate
         the electronic integrator.   Also, record the standard concentration, detec-
         tor attenuation  factor,  chart  speed, sample loop  temperature,  column tem-
         perature  and  identity,  and  the carrier gas type and  flow rate on the data
         form shown in Figure 5-1-  It is also recommended that the same information
         be recorded directly on the  chromatogram.   Record the operating parameters
         for the particular detector being used.
     9.  After the  analysis,  determine the retention time of each standard component
         and determine the peak area. Repeat the injection of the first cylinder gas
         standard  until the area counts from two consecutive injections are within 5
         percent of their average.
    10.  Multiply  the average area count of the consecutive injections by the atten-
         uation factor  to get  the calibration area value for  that standard concen-
         tration.   NOTE:  Attenuation factors which  affect the plot traced,  but not
         the  area count  returned  by  an  electronic   integrator should  not  be
        multiplied by  the  average  area count.   Observe the  effect  of attenuation
         changes   made  at  the  console  of  a   specific  electronic  integrator  to
        determine  the appropriate course of action.

-------
                                                                Section No.  3.16.5
                                                                Date June  30,  1988
                                                                Page 3
                Analysis of Method 18 Field  Samples
Date:   2-/Z /&&  Analyst:  JT 6n>od/i "ok.       Plant:
Location:   PAs/7C . //£ _  Sample Type:
                                                                  C*Ji**  C-
                                                        •/"£.*//<»»-
Type of Calibration Standard:  t-i^meLin-h
Number of  Standards: 3	  Date Prepared:
                                              Target  Compound:
                                             •2-/Z. /3g  Prepared By:
                                                                    A-
                             Column Used: P
                                      4. -Lit
                                                      IZ.4-2.
                                                      J.1.7-
                                                                    A/4-/ AM

                                                                    10 I  /O '
                                                                     2- /  2.
                                                                   £.&3/ J-.&i
                                                                      0.01

                                                                       / 4-2.01.
                                                                      4/6Z
                                                                      /oo
                                                       -1.0
  Linear  regression equation; slope  (m):   4/, BO   y-intercept  (b) :  _
Sample Analysis  Data
 First analysis/second analysis
  Sample identification
  Interface dilution factor
  Flow rate through  loop  (ml/min)
  Liquid injection volume (tubes)
  Injection time (24-hr clock)
  Chart speed  (cm/min)
  Detector attenuation
  Peak retention time
  Peak retention time range
  Peak area
  Peak area x  atten.  factor (At/A2
  Average peak area  value (Y)
  % deviation  from average (#Dav  )
  Calculated concentration (Cs)
                                     Sample  1
                                       A/A
                                     -LOO/ 2J3Q
                                     V4/ A/A-
                                     15:11,11$: W
                                     /O / 10
                                     L- I 2-
                                     ^.g^/^.g3
                                       Q.OI
                                     004-/
                                                    Sample 2

                                                      ygg-2
                                                        A/A
                                                    2oo/ 2,0 Q
                                                    A/A/  A/A-

                                                     rt I  /O
                                                     2- I  7-
                                                                    Sample 3
                                                                    2,00/2.00
                                                                    A/A-/ A/A
                                                                       1
                                                                     Z. 1 2-
                                       H.+Z.
                                                       1*14-0
                                                        0.4-
                                       3^3
                                                       S7.2,
                                                                     4-3.1
              (Y  -  b)
     or
                        *D,
                           avg
                 m
                                                                   _ p
                                                               ltd     act
Figure 5-1.   Data form for analysis  of Method 18 field  samples.

-------
                                                              Section No. 3-16.5
                                                              Date June 30, 1988"
                                                              Page 4

  11.   Repeat the procedure for  the  other  standard concentrations,  with the high-
        est concentration analyzed last.
  12.   Prepare a plot with the standard  concentration  (Cs)  along the abscissa (x-
        axis) versus  the corresponding calibration  area value along  the ordinate
        (y-axis).   Perform a regression analysis  to  calculate  the slope and the y-
        intercept.  Draw the least squares line on the plot.

5.1.2  Preparation and Analysis of Gaseous Standards from High Concentration Cylin-
ders - Gaseous standards can be prepared  from  high concentration  cylinder gases by
dilution with hydrocarbon-free gas and  collection  of the diluted  gas mixture in a
Tedlar bag  (10 liters  or larger).   A  single-stage dilution  system is  used for
dilutions up  to about 20-fold.  For greater dilutions,  a two-stage dilution system
should be used.   It  is  recommended that a check of  the  dilution  system be made by
analyzing a low concentration cylinder standard that is in the range of one of the
 standards prepared  by  dilution.   Prepare  the gaseous  standards, by the dilution
technique, using the procedures that follow:
     1. Assemble  the  single-stage dilution system,  as  shown in  Figure  5-2 and/or
        the  two-stage dilution system, as  shown  in Figure  5-3,   using rotameters
        (flowmeters)   calibrated  following  the  procedures described  in Subsection
        2.1.3-   (More precise  dilutions  may  be possible if  the dilution system
        utilizes  mass  flow  controllers  and  mass  flowmeters   in  place  of  the
        rotameters.)
     2. Connect the primary flowmeters  on the single-stage system to  the calibra-
        tion gas mixture and the  diluent  gas  (hydrocarbon-free).   On the two-stage
        system,  connect the secondary flowmeters to the two diluent gas cylinders.
     3- Connect a leakfree evacuated  Tedlar bag  fitted with  a  quick  connect  or
        equivalent,  compatible to  the  connection on the actual sample  bags or the
        interface sample line, to the tee connector on  the  single-stage system or
        following the second stage of the  two-stage system.
     4. Open  the  gas cylinders,  adjust all  the pressure regulators  to  the  same
        pressure,  and adjust  the  gas flows  to achieve  the desired  dilution ratio
        using the flow control valves.  On the two-stage system,  adjust  the needle
        valve on the  high  concentration waste  outlet so that 30%  of the high  con-
        centration gas is  wasted  and 10% goes  to  the second stage.   NOTE: Divert
        high concentration waste  to  a fume  hood or pass it  through  an appropriate
        adsorption  media  to  protect  personnel  from  exposure' to   harmful
        concentrations of organic  vapors.
     5- Take periodic  readings  of the  pressure difference  between  the  first  and
        second stages of the two-stage system,  as indicated by  a water manometer or
        equivalent,  to..correct the flow reading  from the first  stage to the second
        stage.  If  the  flow rates of  the two stages  can  be balanced so  that the
        pressures  are equal,  then  no  correction will be necessary.
     6. Disconnect the Tedlar  bag  from the dilution system before  the bag is total-
        ly full,  and  turn off  the gases.   Label the bag to  indicate the contents,
        the time  and  data when it  was prepared,  the identity  of the high concentra-
        tion gas  cylinder,  and the dilution ratio(s)  used.
     7- Record the  ambient temperature,  the  flow meter readings,  the  barometric
        pressure,  and the  average first stage pressure  on  the data form  shown  in
        Figure 5•^•
     8. Calculate  the concentration  (Cs ) ,  in ppmv ,  of  the component in  the final
        gas  mixture   by  the  following  formulas  for  single-stage  and  two-stage
        dilution.

-------
COMPONENT
    GAS
 CYLINDER
o
                                              "T" CONNECTOR
                               CALIBRATED ROTAMETERS
                                 WITH FLOW CONTROL
                                       VALVES
    DILUENT
     GAS
   CYLINDER
                                                                             TEDLAR BAG
                                                                                                CO CD

                                                                                                (0 (B
                                                                                                Ul CH
                                                           O
                                                           rt
                                                           H-
                                                                                                 UJ .
                                                                                                 O
                                                                                                 • UJ
                  Figure  5-2.  Single-stage calibration gas dilution
                           system.
                                                        >x> ON
                                                        oo •
                                                        00 U1

-------
                                                       MANOMETER
                                        HIGH
                                   CONCENTRATION
                                       WASTE
                          NEEDLE VALVES
           ROTAMETERS-
                                                                                    LOW
                                                                               CONCENTRATION
                                                                                    GAS
                                              PRESSURE
                                              REGULATOR
+ ( PR
                                 DILUENT AIR
DILUENT AIR
    PURE SUBSTANCE OR
PURE SUBSTANCE/N, MIXTURE
                      Figure 5.3.  Two-stage calibration gas  dilution  system.
                                   T3 O C/3
                                   0 (0 0)
                                   TO rt O
                                   (D (D rt
                                       H-
                                   CT\ C-i O

                                     § 3
                                     (D 2
                                       O
                                    OJ •
                                     o
                                                                                                  co •
                                                                                                  Oovjl

-------
                                                              Section No. 3.16.5
                                                              Date June 30,  1988
                                                              Page 7
          Preparation  of Standards  by Dilution of Gas Cylinder Standards
Date:   2/2-1^
                  Preparer:
                                                Purpose:
Cylinder Component;   /^<^g/?/0  ppm..  Certification Date:   //12. I ft&
          Stage 1
     Standard'gas flowmeter reading
     Diluent gas flowmeter reading
     Laboratory'temperature (°K)
     Barometric pressure (Pb) (mm Hg)
     Flow'rate  of cylinder gas (qcl)
       standard conditions (ml/min)
     Flow rate of diluent gas   (qdl)
       standard conditions (ml/min)
     Calculated concentration (Cs )
Mixture 1
                                                       Mixture 2
                                                          4O
                                                                       Mixture 3
                                      at
                                     at
                                           40
                                                                          /OO
                                        (Xxqcl)
Stage 2 (if used) Mixture 1
Standard gas flowmeter reading
from stage 1
Diluent gas flowmeter reading
into stage 2
Average differential pressure (Pd)
between stage 1 and 2 (mm H20) '
Flow rate of diluted gas
tec2 actual) at standard
conditions to stage 2 (ml/min)
Flow rate of diluted gas
tec 2 corr) at corrected
standard conditions to
stage 2 (ml/min)
/O
10
0
lo
/o
Mixture 2 Mixture 3
2-7- A/-4

0






Flow rate of diluent gas (qd2) at' ' . /
standard conditions to /
stage 2 (ml/min) /& (00 \
Calculated concentration (Cs )

IO.D

30.0 l

                                         actual
                  c.  = x
                              ci + qdl)
                                              ^c 2 corr
Figure
             Calibration standard preparation data form for diluted gas cylinders.

-------
For single-stage dilution:
                                                            Section No. 3.16.5,
                                                            Date June 30,  1988*
                                                            Page 8
                                     (X x qc)
                              Cs = 	                   '    Equation 5"!
   where
             X    = Mole or volume fraction of the organic in the calibration gas
                    that was diluted, ppmv,
             qc   = Flow rate of the calibration gas that was diluted, and
             qd   = Diluent gas flow rate.

For two-stage dilution:

                        Sc2 corr,3 ^c 2 actual  X 	             Equation 5~2


   where

   qc2 corr     = Corrected flow rate from the first stage to the second stage,
   qc2 actual   = Actual flow rate from the first stage to the second stage,
   Pd           = Average differential pressure  between the first and  second
                 stage, mm or in. H20, and
   Pb           = Barometric pressure, mm or in.  H20.

   and
                          "c 1           ^c 2  c o r r
   Cs           = X x  	 x 	:	                Equation 5-3
                      (qcl  + qdl)   (qc2 corr  +  9d2)
   where

    X           =  Mole or volume  fraction of  the organic in  the calibration gas
                  that was diluted,
    qcl         =  Flow rate of the calibration gas diluted in the first stage,
    qdl         =  Flow rate of the diluent gas in the first stage,'and
    qd2         =  Flow rate of the diluent gas in the second stage.

   9-  Prepare two more  calibration standards  from the high  concentration cylin-
      der gas  sufficient  to bracket the expected  concentration in  the source
      samples.
  10.  Analyze  the  calibration  standards  following  the  procedures described in
      Subsection 5-1-1  for  commercial gas cylinder mixtures.  Use  a  pump on the
      outlet side of  the  sample loop to flush the standards  through  the loop at
      100 cc/min  for  30 seconds.   Using '-a manometer  connected  to a  tee on the
      outlet of  the  sample loop,  make  certain  that the  sample  loop  pressure
      during analysis of the calibration  standards is  equal  to  the loop pressure
      during actual sample analysis.
  11.  Once the calibration curve is established, it is recommended,  if available,
      that an undiluted cylinder  standard in  the range of the standard  curve be
      analyzed to verify the dilution ratio.  Analyze  the cylinder and calculate
      the sample  area  value by  multiplying  the  peak area  by  the  attenuation

-------
                                                              Section No.  3-16.5
                                                              Date June 30, 1988
                                                              Page 9

        factor.   Use the slope and  the y-intercept  derived from the linear regres-
        sion equation and  the  sample area value to calculate  the cylinder concen-
        tration (Cs)  by the following formula:

                                         Y - b
                                    Cs = 	                   Equation 5~l
     where
                  Y = Sample peak area, area counts,
                  b = y-intercept of the calibration curve, area counts, and
                  S = Slope of the calibration curve, area counts /ppmv .

The  calculated  concentration  of  the  undiluted  cylinder standard,  based  on  the
analysis and the calibration curve generated  from the diluted standards, should be
within 10% of the true value of the undiluted cylinder.  If this criteria cannot be
met, then the GC  calibration  should be checked,  the  diluted  sample may be outside
the  calibration  range,  or there  is  a problem  with  the  dilution system  used to
prepare the standards  (e.g. the rotameters are out of calibration, etc.).  Identify
the problem and correct  it, or use one of the other approaches for preparing cali-
bration standards.

5.1.3   Preparation  of Calibration Standards by Direct  Gas Injection -  This proce-
dure is  applicable  to organic compounds  that exist  entirely as  a  gas at ambient
conditions.  The standards are prepared by direct injection of a known quantity of
a "pure" gas standard into a 10-liter Tedlar bag containing 5-0 liters of hydrocar-
bon-free air or nitrogen.   If  there is more than one target compound then multiple
component standards  can  be prepared by this  method  provided  the relative elution
pattern  for  the  compounds  is  known for the  GC  column being used.   The following
procedures are used to prepare standards by direct gas injection:
     1. Evacuate  a  previously  leak checked,  leakfree  10-liter  Tedlar  bag (also
        checked for  zero  retention)  equipped with  a quick connect  or equivalent
        compatible to the  connection on the Tedlar bag or  the interface sample line
        and preferably fitted with a septum-capped tee at  the bag inlet  (see Figure
        5-5).
     2. Fit  a  septum  to the   outlet  of the  gas  cylinder containing  the standard
        component .
     3- Meter 5-0 liters of hydrocarbon- free air or nitrogen into the bag at a rate
        of 0.5 liter/min using a dry gas meter that has been calibrated in a manner
        consistent with  the procedure described in Subsection 2.1.2.  At the start,
        record dry gas meter pressure and temperature.
     4. While the bag is  filling,  fill and purge a  0.5-ml gas-tight syringe with
        the  standard gas  by   withdrawing  the gas  from  the  cylinder  through  the
        septum.   Repeat  the .fill and purge of the syringe seven times before final-
        ly filling the syringe  and capping  the needle. with a  GC septum.  Allow the
        syringe temperature, to equilibrate with the ambient temperature.
     5. Immediately  before injecting  the  gas  into  the  bag  through  the  septum,
        remove the  septum  cap,  and adjust the  syringe  to  the desired  volume by
        expelling the  excess gas.   The syringe should now be  at  ambient pressure.
        Inject the gas into the bag through the septum (through the side of the bag
        if it has not  been fitted with a septum),  withdraw the syringe, and imme-
        diately cover any  resulting hole with a piece of masking tape or the equi-
        valent.

-------
                                                                                        Septum
                                                                                        Tedlar Bag
                                                                                        Capacity
                                                                                        10 Liters
                                  Dry Gas Meter
            Nitrogen
            Cylinder
no O C/D
(B PJ 0>

-------
                                                              Section No. 3.16-5
                                                              Date June 30, 1988
                                                              Page 11

     6. Record  the final  dry gas  meter  temperature  and  pressure,  turn  off the
        dilution  gas,   and  disconnect  the  Tedlar  bag;  record  the  ambient
        temperature and  pressure on a  data form such  as the one shown in Figure
        5.6.
     7- Place  the bag  on  a smooth  surface,   and  alternatively depress opposite
        sides of the bag 50 times to mix the gases in the bag.
     8. Calculate  the organic standard  concentration  in the bag (Cs) in ppmv using
        the following formula.

                             '     293 x P,                 Ps x Tm
                       G. x IO6  x 	      G.. x IO3 x 	
                                  Ts x 760                 Ts x Pm
                Cs = -  = -    Equation 5~5
                              293 x P •                v  x y
                                ~^    n>     _           fn
                     v  x y x - x io3
                              T  x 760

where
                G   = Gas volume of  organic  compound injected into the Tedlar bag,
                      ml.
               IO6  = Conversion to ppmv, ul/liter,
               PS   = Absolute pressure of syringe before injection, mm Hg,
               Ts   = Absolute temperature of the syringe before injection, °K,
               Vm   = Gas volume indicated by dry gas meter, liters,
               ?    = Dry gas meter correction factor, dimensionless,
               Pm   = Average absolute pressure of the dry gas meter, mm Hg,
               Tm   = Average absolute temperature of dry gas meter, °K, and
               IO3  = Conversion factor, ml/L.

        Note: The  syringe pressure  and absolute  temperature should equal the baro-
        metric pressure and the absolute ambient temperature.
     9- Prepare  two  more calibration standards sufficient  to  bracket the expected
        concentration in the source samples.
    10. Analyze  the  calibration  standards  following  the procedures  described in
        Subsection 5-1-1  for  commercial  gas cylinder mixtures.  Use  a pump on the
        outlet side of  the  sample loop to flush the standards  through the loop at
        100  cc/min for 30 seconds.   Using  a manometer  connected  to a  tee  on the
        outlet  of the  sample loop,  make certain  that  the sample  loop  pressure
        during analysis of the calibration standards is  equal to the loop pressure
        during actual sample analysis.

5.1.4  Preparation of Calibration Standards by Liquid Injection - This procedure is
used to  prepare gaseous  standards  in Tedlar bags  from  liquid  organic  compounds.
The liquid compounds  used must be 99-9% mole pure  or the purity must be  known to
calculate the gaseous  standard concentrations.   If  there is more  than  one target
compound, then multiple component standards can be prepared by this method provided
the relative  elution  pattern for the compounds  is  known for the  GC  column  being
used.   Use the procedure that follows to prepare  standards by this  technique.
     1. Assemble the equipment shown in Figure 5-7 using a dry gas  meter calibrated
        following the procedure described in Subsection 2.1.2 and a water manometer
        for the pressure gauge.  All  connections should  be  glass,  Teflon,  brass or
        stainless steel with quick connects or equivalent, compatible to the con-

-------
                                                               Section No. 3.16.5
                                                               Date June 30, 1988
                                                               Page 12

        Preparation of Standards  in Tedlar  Bags  by Gas  and Liquid Injection
Date:
Preparer: >77-
','ck
Purpose:
he Id
Organic Compound: rc^o^/./T^r^&'H^Le^i^
Compound Source: Fi^kt*- '• Compound Purity (P) : f?i

Gas: or Liquid: «-^''
t-% Compound Mole Weight (M) : /^iTg


Gas Injecti6n Mixture 1 Mixture 2 Mixture 3
Bag number or identification A// A
A//A A/
/A
Dry gas meter calibration factor (Y) . ' \ ' ' \ \
Final gas meter reading, liters / ) /
Initial -gas meter reading, liters /
Volume metered (Vm ) , liters
Ambient temperature, °C
Average gas meter temperature, °C
Absolute' gas meter temp. ; (T.m.) ,• °K
Barometric pressure (Pb ) , mm Hg
Average gas meter pressure, , mm Hg
Absolute gas meter press. (Pm)<, mm Hg
Gas volume injected (Gv), ml
Syringe -temperature (Tg ) , °K
Absolute, syringe pressure (P ) , mm Hg
Calculated concentration (Cs )

P x T
s rn
("!' V 1 0^ M
uv x iu x
*T v D
N 1s X ^m
c - -
Us
V x Y
Liquid Injection Mixture 1
Bag number or identification S - 1
Dry gas .meter calibration factor (Y) 
Liquid volume injected (Lv), ul 2..OO
Calculated concentration (C ) 3-1, 7-

P ^ i/i v i n* v
°s b"-' X 1U "
M x V x Y x P
/ 1
(

\
\ \




}
/r
\

Q
s c a 1 c .
C - x 10
~s c o r r . " "w
P




\





I

Q%
Mixture 2 Mixture 3
3-Z S
O.H-3>O O.
-3
743£
38.10 74-.S2.
3t.10 2,0. ?0
7.00 $S.<*2-
Z-6 2-6
30/ ^
23
-?>&.Q' 7S-6.4
1.4- /•
75-1-6. 7t
f
O.3
/.kZ3 /, 623
2..00 ;, ;. z
00
7^..*;' \ /4--5~

Cs calc.
r - 	 v 1 n
Us corr. - X 1U
P

Q%
Figure 5.6.   Calibration data  form  for preparation of standards  in Tedlar bags by
             gas and liquid injection.

-------
                                                       SYRINGE
                                                          11
                                             BOILING
                                             WATER
                                              BATH
                                                            -^-SEPTUM


                                                            MIDGET IMPINGER
                               ETER
NITROGEN CYLINDER
                                                                         TEDLAR BAG
                                                                           CAPACITY
                                                                           50 LITERS
              Figure 5-7.  Apparatus for preparation of calibration standards  by  liquid
                           injection.
no o co
  » 0)
  ft O
  (D rr
H- «-( O
OJ C 3

  (0 -Z
    o
  oo .
  O
  -  OJ
                                                                                                  vo o\
                                                                                                  00 •
                                                                                                  OOU1

-------
                                                          Section No. 3.16.5
                                                          Date June 30, 1988
                                                          Page 14

      nection on the Tedlar bag or the interface sample line, for connection to
      the Tedlar bag.
   2. Allow  the  liquid organic  compound to  come  to ambient  temperature, and
      determine  the  density of  the  liquid by  weighing the liquid  in  a  tare-
      weighed  ground-glass stoppered  25~ml  volumetric flask  or ground-glass
      stoppered  specific  gravity bottle.   Calculate  the  density in  terms of
      g/ml.   As  an alternative, use  a literature value of  the density of the
      liquid at 20 °C.
   3.. Leak check  the  system by pressurizing it to 5 to  10  cm  (2 to 4 in.) H20
      and shutting off  the diluent gas supply.  The system is leakfree if  there
      is no  change in  the pressure after  30 seconds.   If the  leak  check is
      good,  release  the pressure.   If the  system  fails the leak check, locate
      the leak using a soap solution and repair the leak.
   4. Connect  a  quick connect or  equivalent, compatible to  the connection on
      the Tedlar bag or the interface sample  line, to a leakfree uncontaminated
      50-liter Tedlar bag.  Evacuate the bag.
   5. Turn on the hot plate and bring the water to a boil. :
   6. Connect the bag to the impinger outlet.
   7. Record  the initial  meter  reading and  temperature.    Open  the  diluent
      supply valve,  and adjust the flow rate to about  3 liters/minute so  that
      the bag  will fill in about 15  minutes.  .Record  the  meter pressure and
      temperature  and  the barometric pressure  at  the  start on  a form  such as
      the one shown in Figure 5-5-
   8. Use  a  clean 1.0-  or 10-microliter syringe  with a needle of sufficient
      length to inject  the liquid below  the air inlet  branch of the tee on the
      midget impinger.
   9. Fill  the syringe  to the  desired volume with  the  organic  liquid, and
      inject the  liquid by inserting  the  needle  through the  septum until the
      needle is  below the  air inlet.   Depress the  syringe plunger completely
      over a period  of about 10  seconds and withdraw  the  needle.   NOTE:  When
      dispensing liquid from a syringe,  take  care  to account for the volume of
      liquid present  in the syringe needle.   In general,  the  potential  error
      resulting from  the volume  of the needle  is  most  conveniently avoided by
      ensuring  that  the   needle  volume  is  completely  full   of liquid  upon
      filling the syringe  and dispensing from it.   If  air pockets exist in the
     .syringe after filling, this will be almost impossible.
  10. When the bag is almost filled,  record the water  manometer pressure.  Turn
      off the diluent gas  supply,  and disconnect the bag.   To  equilibrate the
      contents in the bag,  either set  the bag aside  for an hour or massage the
      bag by alternately depressing opposite sides of the bag 50 times.
  11. Record  the  final meter  reading  and  temperature.   Calculate  the  con-
      centration of the calibration standard  (Cs)  in the bag in ppmv  using the
      following formula.

      Lv
      — x p x 24.055 x 106
      M                                       Lv x p x Tm
Cs = 	 =  6.24 x 10A x 	     Equation 5-6
              293 x Pm                      M x Vm  x y x Pm
     V  x Y x 	 x 103
              Tm  x 760

-------
                                                              Section No. 3-16.5
                                                              Date June 30, 1988
                                                              Page 15

    where
             V      = Gas volume indicated by dry gas meter, liters,
             L      = Volume of liquid organic injected, ul,
             M      = Molecular weight of the organic compound, g/g-mole,
             p      = Organic liquid density, g/ml,
             24.055 = Ideal  molar  gas volume  at  293 °K  and 760 mm  tig,  liters/g-
                      mole,
             106    = Conversion to ppmv, ul/liter,
             103    = Conversion factor, ul/ml, and
             y      = Conversion factor for dry gas meter.

    12. When using a liquid standard  that  is not  99-9$ pure, use  the following
        formula to correct the calculated concentration of the calibration standard
        (Cs corr) in PPmv.

                                       Cs calc
                             Cs corr = 	 x P                   Equation 5-7
                                        100%
     where

             Cs corr  = Corrected calibration standard concentration, ppmy ,
             Cs calc  = Calculated  calibration standard  concentration  (Cs),  ppmv ,
                        and
             P        = Purity of liquid organic compound, percent.

    13. Prepare  two  more calibration standards sufficient  to bracket the expected
        concentration in  the source samples.
    14. Analyze  the calibration  standards  following  the procedures  described in
        Subsection 5-1-1  for commercial gas cylinder mixtures.   Use a pump on the
        outlet  side  of  the sample loop to flush  the standards through the loop at
        100  cc/min for  30 seconds.   Make  certain  that  the  sample  loop pressure
        during  analysis  of the  calibration standards is equal to the loop pressure
        during  actual sample analysis.
    An  alternative procedure, subject  to the  approval of  the  Administrator,  for
preparing gaseous  standards  from  liquid organics  substitutes a heated GC injection
port for the midget impinger setup and, due  to the high back pressure of the injec-
tion port, a calibrated mass flowmeter  for the dry gas meter.

5.1.5   Development of  Relative Response  Factors  and Relative  Retention Factors-
For emission tests where on-site GC analysis  involving more than one organic com-
pound  will be  conducted,  the development and use  of relative response factors and
relative retention times is  recommended.   In the  laboratory,  gaseous calibration
standards are prepared  for each  target  organic compound and analyzed by one of the
techniques described  in the previous subsections.   Choose one  of  the target com-
pounds  or  prepare and generate another calibration curve for  a different organic
compound to be used to calculate the relative response factors and retention times.
The compound selected  should exhibit  a  retention  time  comparable  to  the  other
target  compounds,  should be stable, and/or  easy  to prepare and use  in the field.
This procedure must be verified in the laboratory prior to field testing.
    The relative  response  factors  are calculated  by dividing  the slopes of  the
target  compound calibration  curve by  the  slope of the selected organic calibration
curve.  The y-intercept  from the  regression  equation is ignored in calculating the
relative  response factors.   It  should  be  noted  that  a  very large  y-intercept

-------
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 16

(greater than 5%  of  the slope)  for any compound may  adversely affect the validity
of  this  calibration  technique.   During  analysis  of  field samples,  the selected
organic  compound  can  be used  to calibrate  the  GC  detector  response  and  column
performance.  The response  factor determined  in the  field for  the selected organic
is  used  to calculate  the  field  response factors for  the other  target compounds
using the  relative  response factors determined previously  in  the laboratory.   The
same approach  is  used,  to  predict the retention times  of  target  compounds  in the
field using the selected compound retention  time  determined in  the  field and the
relative retention times for the  target compounds determined in the laboratory. Use
the following procedures to develop relative response  factors  and relative reten-
tion times.
     1. Generate,  at  the minimum, a three-point calibration curve for each target
        organic compound using gaseous standards following  the procedures described
        in the preceding subsections.  Record the retention time of each compound.
     2. Select  one  of the  target compounds,  preferably with a  retention time be-
        tween the other target compounds,  or generate  another  calibration curve,
        with a  minimum of  three  points,  for  a non- target  organic compound  with a
        comparable retention time.  Select  the standard compound to be used in the
        field based primarily on  the ease of  use.   Determine the retention time of
        the selected compound (if not already determined) .  Measure the carrier gas
        flow rate using a bubble-type flowmeter or other suitable  flowmeter.
     3- Inject a sample of  the diluent gas,  and determine the retention time of the
        unretained diluent  peak.   This is needed to  calculate  the relative reten-
        tion by the following formula:
          rx/s = -                                         Equation 5-8
                  (tRsi - fcMi)

     where

          rx/s = Relative retention time based on adjusted retention volumes of the
                 target compounds and the selected compound, cc/cc,
          tRxi = Initial retention time of compound x, seconds,
          tMi   = Initial retention  time of  unretained  diluent gas  peak,  seconds,
                 and
          tRsl = Initial retention time of selected organic compound, seconds.

     4 .  Calculate the relative response factor for each target compound relative to
        the compound selected in step 2 using the following formula.

                   ms
             FRx = —                .                                  Equation 5-9
                   mx
     where

             FRx = Relative response factor for compound x, dimensionless ,
             ms  = Slope from the  calibration  curve regression  equation  for the
                   selected organic compound, area counts/ppmv , and
             mx  = Slope from the calibration curve regression equation for
                   compound x, area counts/ppm .

-------
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 17

     5.  To verify  that  the  relative  response  factors  are  correct,  simulate  the
      '  transportation of the  GC  to the field by turning off  the detector,  the GC
        oven and  the carrier  gas  flow overnight or  longer.   After  the simulated
        period has elapsed, turn on  the GC,  the  carrier gas,  and the detector,  and
        establish the analytical conditions that were used to generate the relative
      '  response factors.  Measure the carrier gas flow rate.       .   ...  ..  ..-
     6:  Recalibrate the GC by  generating  a three-point calibration curve using the
        selected organic compound,  analyze  each  of  the target compound calibration
        standards, and a diluent  gas sample.  Calculate  the  concentration of each
        target compound using the relative response factor for the compound and the
        slope from  the new calibration curve determined for  the selected organic
        compound with the following formula.

                      Yx
                C  = 	x FD                                        Equation 5~10
                 X          K X
                     ID
     where

             C     =  Calculated concentration of  compound  x calibration  standard,
             Yx    =  Detector  response  for compound x  calibration standard, area
                      counts,
             ms t d  =  Slope  from  new  calibration  curve  generated  for  selected
                      organic standard compound, area counts/ppmv,  and
             FRx   =  Relative response factor for compound x, dimensionless.
        The calculated value for each target compound using the compound's relative
        response factor must be within  5% of the actual standard concentration for
       • this technique to be used for that compound.
     7. Determine  the predicted  retention  times  for  the target  compounds using
        their  relative  retention times  and  the retention  time  determined for the
        selected organic compound using the  following formula:

               rrxf = ((rTsf - tMf ) x rx/s) + fcMf                       Equation 5-11

     where

               rTxf = Calculated retention  time for compound  x  using the relative
                      retention time factor, seconds,
               rTsf = Measured  retention  for  selected  organic  compound   during
                      second analysis, seconds, and
               tMf  = Measured retention time of unretained diluent  gas peak  during
                      second analysis, seconds.
        The calculated retention time for the  target compounds should agree  within
      •  one second or 2%,  whichever is  greater, of  the  actual retention time seen
        for the target compounds during the second analysis.
     8. Record all data on a form such as the one shown in Figure 5-8.

5.1.6  Calibration Standards for Adsorption Tube  Samples  -  The  calibration stan-
dards necessary  for  the  analysis of adsorption tube samples  differ from the stan-
dards described  in  the  previous subsections in  that  the adsorption tube standards
are liquid rather  than  gaseous.   The liquid standards can be prepared directly in
the desorption  solvent following  the procedures  described in the methods   refer-
enced in Table F or,  subject to the approval of the Administrator, on blank adsorp-

-------
                                                        Section No. 3.16.5
                                                        Date June 30, 1988
                                                        Page 18

Development of Relative Response Factors and Relative Retention Factors
Date: 2-/2~b /&& Preparer: \7- 0?oo
Target Compound: Perchloroe'ttwitA**?-
Surrogate Compound: -Lsobnifje-

Target Compound Calibration Data
First analysis/verify analysis
Standard concentration
Flow rate through loop (ml/min)
Liquid injection volume (tubes)
Injection time (24-hr clock)
Chart speed
Detector attenuation
Peak retention time (tRxi/tRxf)
Peak retention time range
Peak area
Peak area x atten. factor (Yi/Yx
Verification analysis conc.(Cx)
Percent deviation from actual
Calculated retention time (rTxf)
Percent deviation from actual
Linear regression equation; slope
Surrogate Calibration Data
First analysis/second analysis
Standard concentration
Flow rate through loop (ml/min)
Liquid injection volume (tubes)
Injection time (24-hr clock)
Chart speed
Detector attenuation
Peak retention time (tRsi/tRsf)
Peak retention time range
Peak area
Peak area x attenuation factor
dri £^ Purpose
Type of Standard
Type of Standard
Standard 1
1.1* Pf"
2_c>0 / tot
V/f- / /W
;4:0o//4:3'
/Ocml*'*l ID
Z, 1 -Z-
1.43/ 2.4-S
O.02-
3^3 / 3.4^
) 74(, / 6,12.
1.14
— / 4 *7
(/« 6* fO
A-.4-Z
—O.(j>7<>
(mj: 73.48

Standard 1
/*.*//*
JLOO/ 2-00
f/A-l ti*
I4:oi/ I4.\2 1 llt^
Linear regression equation; slope (mo ) : ^ (m,):

Nonretained peak retention time (tMi

/tMf): O.tl 10.

'. die. Cotrtrr\jtr-i
: t-iqmjl ,'» ,
'. O<*S Cy It'n.
'
Standard 2
f,.*
2JOO /zoo
A/A-/ At 'A
/$:(>$/ /V4O
10 I 10
2- 1 Z-
Z. 44/^.4-5
(?.<53
/^/2/ /?6>3
3&Z^/ 3924
^"2.4
V-/.4
.2.42-
0. (*7o
y-intercept
Standard 2
SO-O^M
^roo/zx>o
AjA-tfJA
/$: f1/ /S-IZ
10 / 10
1 1 1
3.2^/3-2.6
O.OI
o^ 7/f / / *» *uG ' A~.
^^ *\0 \O / ^**i1 'i
j-f (f
    y-int 43 Relative Response Factor (FRjc): /. 3-J- Relative Retention Factor sML*u*.>'»U""i'xA* B^ di£r Standard 3 zool zoo A/A-/ A/A 14 '.ft/ 14-:£& 10 //O 2-1 2- 2.4-4/^.4-2- O.O2- 2>(e52./ $(?(*<<> 730^/7332- 97-. / -1.4- 2.4--Z- C>.6>% (b) : !C>.1(f Standard 3 /(X>fP«\ ^00/^,00 /VA 1 A/A /5:4?/ ' Ib'.Cfl /O 1 (t> 2-1 2. Z.ll 1 5. lie, O-OI 4?\(*/ 4152- 1437.1 1S04 ercept (b) : -/g.fW / „ ) . £). 3L \LK/ 5 1 ' ms m -x/s °Mi s t d x F Rx "Mf Figure 5.8. Data form for development of relative response and relative retention factors.

-------
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 19

tion tubes and then desorbed.  Both  methods  require  similar preparation and analy-
sis of  standards and  desorption  efficiency samples,  but  the way  the calibration
curve is generated is different.
    For  calibration  standards prepared directly  in  the  desorption  solvent,  the
standards are used to generate the calibration curve, and the desorption efficiency
is  determined  separately.    The  calculated desorption  efficiency is  then  used to
correct the analytical results for the  emission  test samples.   The disadvantage of
this method is that  the  desorption efficiency may not  be  constant for each level.
This can result  from a constant amount  of  analyte being retained by the adsorbent,
instead of an amount proportional to the total amount of analyte on the adsorbent.
When using the desorption efficiency to correct each analytical result, the analyst
must use  the  desorption  efficiency determined for the concentration level closest
to that of the sample.
    For  calibration  standards prepared on absorbent material,  the  desorbed solu-
tions are used to generate the calibration curve.   By this procedure, the desorp-
tion efficiency  is already taken into account when calculating the organic compound
catch of  the  adsorption tube  samples.   Liquid calibration  standards  must also be
prepared  to calibrate  the  GC to determine if  the desorption efficiency is greater
than 50%-   The  advantage  of this  method is that  both level-dependent or absolute
amounts  of  organic  compounds not  desorbed from  the  adsorbent  are  automatically
taken into consideration.
     For  maximum accuracy,  preparation of  standards  directly  in  the desorption
liquid or on adsorbent will  require  the preparation of  a relatively large volume of
a  high  concentration  working standard  from which  the calibration  standards are
prepared.  The working standard should  be  100  times  more concentrated than the
highest concentration  calibration standard.  Three levels of calibration standards
should be prepared to bracket the  expected concentration  of  the liquid resulting
from desorption  of  actual samples.    The concentration  of the  sample desorption
liquid will depend on the catch weight of the target  organic  compound(s)  and the
amount of desorption liquid used (1.0  ml  per  100 mg  of  adsorbent material).   The
catch weight  will in  turn depend  on the  sample  volume of  flue  or  duct gas drawn
through the tubes and  the  concentration of the emission source.   Use  the following
formula to estimate  the  concentration (Cs),  in ug/ml,  of the midrange liquid stan-
dard that will be approximately equal to the actual samples:

                                      V.d X Cc X M
                                 Cs  =	                   Equation 5-12
                                       24.055 x Ld
where

     V  ed  = Predicted gas sample volume,  liters,
     Cc      = Concentration of the organic compound at  the source, ppmv
              (ug-moles/g-mole),
     M      = Molecular weight of organic compound, ug/ug-mole,
     24.055 = Ideal molar gas volume at 293 °K and 760 mm Eg, liters/g-mole, and
     Ld      = Volume of desorption liquid,  ml.

The preliminary  survey  sample  results should  be used to  calculate  the  required
calibration standard concentrations.
    To prepare adsorption tube standards,  use the procedure described in the refer-
enced method  or  the alternative  procedure,  subject  to the prior approval  of the
Administrator.   Regardless of which  type of calibration  standard is selected, use
the following procedures to prepare the standards:

-------
                                                              Section No. 3-16.5
                                                              Date June 30, 1988
                                                              Page 20

     1. Prepare a working  standard  for  each organic compound by weighing each com-
        pound into an individual tare-weighed ground-glass stoppered 250-ml volume-
        tric flask.  Dissolve  the  compound in the proper desorption solvent speci-
        fied by the referenced  method in  Table E of the Method Highlights Section.
        Dilute the  solution  to volume.   Refrigerate  the  working standard when not
        being used.                                               '
     2. Using a clean  microliter syringe, transfer the required amount of working
        standard to a vial equipped with  a Teflon-lined septum top,  and add suffi-
        cient desorption solvent to achieve a final volume  equal  to  the volume of
        desorption  solvent required for  actual  samples.    Cap  the vial  with the
        top, shake the vial to mix  the contents.  NOTE: When dispensing liquid from
        a syringe,  take care  to account  for  the  volume  of liquid present  in the
        syringe needle.  In general, the  potential error resulting from the volume
        of  the  needle  is  most  conveniently avoided by  ensuring that  the  needle
        volume is completely full of liquid upon filling the syringe and dispensing
        from it.   If air pockets exist in  the  syringe  after filling,  this will be
        almost impossible.
     3. Establish the optimum  GC conditions determined during  the  analysis  of the
       • preliminary survey samples.
     4. Select a  suitably sized injection  syringe (5~ or  10-ul),  and  flush the
        syringe with  acetone  (or  some other suitable solvent  if acetone  is the
        standard component) to clean the syringe.
     5. Flush the  syringe  with standard solution  by  withdrawing a syringe full of
        the  solution  from  the  septum  vial,  and dispensing  the  solution  into  a
        beaker containing charcoal  adsorbent.
     6. Refill the  syringe with standard  solution, withdraw the  syringe from the
        vial, and wipe the syringe needle with a laboratory  tissue.
     7. Adjust the  syringe volume  down  to the desired  amount  (see NOTE under Step
        2) ,  and inject  into  the GC.  Note the time of the  injection  on the strip
        chart recorder and/or  actuate the electronic integrator.  Also, record the
        standard concentration,  detector attenuation factor, chart speed, injection
        port temperature, column temperature and identity, and the carrier gas type
        and flow rate on the form shown in Figure 5-9-  It is also recommended that
        the same information be recorded  directly on the chromatogram.   Record the
        operating parameters for the particular detector being used.
     8. After the analysis, determine the retention time of the standard component
        and  determine  the peak area.  Repeat  the injection of the  first  liquid
        standard until  the area count  from two consecutive injections  yield area
        counts within 5 percent of  their average.
     9. Multiply the average area count of the consecutive injections by the atten-
        uation factor to get  the calibration area  value  for that standard concen-
        tration.
    10. Repeat the procedure for the other standard concentrations.
    11. Prepare a plot with the standard  concentration  (Cs)  along the abscissa (x-
        axis) versus the corresponding  calibration area values  along  the ordinate
        (y-axis).   Perform a regression analysis  to calculate  the  slope and  the y-
        intercept.   Draw the least squares line on the plot.
    To  determine  the  desorption  efficiency  for  the  target  organic  compound(s)
requires  spiking  the  target organic compound(s)  onto  the  absorbent material and
desorbing the compound (s)  using the same  procedures  that will be used  for  actual
samples;  the desorption  solution is then  analyzed.  The spikes  should  be prepared
at three  levels in  the  range  of the source  samples.  The  following procedures are
used to determine the  desorption efficiency:

-------
                                                              Section No.  3.16.5
                                                              Date June 30,  1988
                                                              Page 21

     1.  Place an  amount  of adsorbent  material  equivalent to  the amount used  for
        actual tube samples in a vial with a Teflon-lined  septum  cap.   Prepare  ten
        vials (three sets of triplicates and  one blank).
     2.  Using a clean microliter  syringe,  aliquot from the working  standard solu-
        tion, in  triplicate  into  each  set of  vials, an  amount of spike equal to
        each level of calibration  standard.
     3.  Cap each vial immediately  after spiking,  and- allow the vials  to  sit undi-
        sturbed for the 30 minutes.
     4.  To desorb  the  spiked  organic compound(s), dispense the appropriate volume
        of desorbent  solvent  and   treat  the  vials as  specified by  the  referenced
        method (Table E).  Prepare a blank vial  containing adsorbent  and desorption
        solvent only.
     5.  Analyze the desorption solutions following steps  4 through 8  used above  for
        the  calibration  standards.   Record  the data on  the  form shown  in Figure
        5-9.
     6.  Multiply the average area  count of the  consecutive injections by the atten-
        uation factor to  get  the  area  value for  that sample.  NOTE:   Attenuation
        factors which affect  the  plot  traced,   but not the area  count returned by
        an electronic integrator  should not  be multiplied  by  the  average  area
        count.  Observe the effect of attenuation changes  made  at the console of a
        specific  electronic  integrator to  determine the appropriate  course  of
        action.
     7.  If the  desorption solutions  are to be used  to  generate the calibration
        curve, then  plot  the  expected standard solution concentrations  on  the  ab-
        scissa  (x-axis)  and  corresponding  area  value on  the ordinate  (y-axis).
        Perform a  regression  analysis and draw  the least  squares  line on the plot.
        NOTE:   If the desorption  efficiencies  of the selected  solvent  vary with
        concentration  for any of  the  organics to  be  analyzed,   the relationship
        between the expected standard solution concentrations and the corresponding
        area  value will not  be  strictly linear.    Evaluate  the  linearity  of  the
        resulting plot using  control samples, and obtain  the  prior approval of  the
        Administrator before  utilizing  a  least squares  line generated  from such
        data.
    8.  Calculate  the desorption  efficiency  (DE),  in percent,  for  each level of
        spike using  the calibration area  for  the corresponding  standard prepared
        directly in the desorption solvent using the following formula:

                                    As  - Ab
                               DE =	x 100%                      Equation 5-13
                                      AC .
  where

          AS  = Average area value  for desorption carried  out at given concentration
               level, area counts,
          Ad  = Average area value  for desorption  carried  out  on blank sample, area
               counts, and
          AC  = Average calibration area  value for the corresponding  standard level
               prepared directly in the desorption solvent, area counts.

     The desorption efficiency achieved  at each level must be greater than  $0%  for
the adsorption tube sampling and analytical  method to be  acceptable.   If adsorption
tubes have become the only remaining sampling option, and the  50%  criteria cannot

-------
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 22

         Preparation of Liquid Standards and Desorption Efficiency Samples
Date:   Z/23/68   Preparer:   xT".
         *
                                               Purpose:
Organic Compound:
Compound Source:
Adsorbent Material:
                                                        Gas:
                                                                   or Liquid:
                                      _      _
                         Compound Purity (P) :  f*?f- % Compound Mole Weight- (M) :
                                   Batch No:  /^o    Desorption Solvent: £//•,
Standards in Solvent
Desorption solvent volume (Vs),  ml
Compound spike amount (Vo),  ul
Organic compound density (p), ug/ul   /. fc2-S
Standard concentration (C ), ug/ml
                                      Mixture 1
                                         4.00
                                         4.00
                                                    Mixture 2
                                                      R.OO
                                                       3.
                                                                  Mixture 3
                                                                    4-. 00
                                                                   /Z. 00
Standards on Adsorbent Mixture 1
Adsorbent amount, g
Compound spike amount (V0), ul
Organic compound density (p) , ug/ul
Desorption solvent volume (Vs ) , ml
Desorption time, min.
Standard concentration (Cs ) , ug/ml

6.&DD
A-.OO
/, 0>23
4. no
30
/•t>2-

Mixture 2
0-800
6.00
1, (p2"b
4-. oo
30
3.E5"

Mixture 3
0-800
/3-.OO
1 . 6>23
•4. 00
5O
^ 4«f

Blank
Q.60D
4.00
O

  GC Operating Conditions
  Injection port temperature,  .°C
  Carrier gas flow rate,  ml/min
  Column temperature:
       Initial,  °C
       Program rate,  °C/min
       Final, °C
                                       70 -A
                                      / CO
                                          P^
Chromatographic Results
Injection time, 24-hr clock
Distance to peak, cm
Chart speed, cm/min
Retention time, min
Attenuation factor
Standards in desorption solvent:
     Peak area (Ac),  area counts
Standards and blank from
  adsorbent material:
     Peak area (As and Ab),
        area counts
                                      Mixture 1   Mixture 2
                                        H: ^(*      /5-'04-
                                                              Mixture 3  Blank
                                                                /5V 5?
                                                      /£>
                                                                 /O
                                                                          /O
                                         4.7^
                                                     4.73
                                                      3-
                                                     701
                                                    544
  Desorption Efficiency Calculation
  Desorption Efficiency (DE),  %
                                        Mixture  1    Mixture 2    Mixture  3
                                          g2.8
                                                       77.
      'Vo x p x P
  C  =
                                               As  - Ab
                                           DE  = 	 x  100#
       Vs  x 100%
Figure 5-9-   Data  form  for  preparation  of  liquid  standards  and  desorption
             efficiency samples for adsorption tube analysis.

-------
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 23

be met,  then, subject  to the  prior approval of  the Administrator,  explore more
vigorous desorption techniques such as longer desorption times, sonification of the
vials during desorption, and/or other desorption solvents.

5.2  Audit Sample Analysis

    After analysis  of the calibration  standards,  and generation  of a calibration
curve,  conduct  the analysis  of the  audit cylinder(s).   Audit samples  should be
introduced into the GC by the same procedure used for the calibration standards. If
possible, the audit sample  should  be introduced into the probe for the direct and
dilution interface techniques.  The  audit  sample  analysis must agree within 10$ of
the actual concentration of the audit sample  before sample analysis can begin.  If
the audit  criteria is not  met,  first try  recalibrating the GC with the existing
standards,  and  then  reanalyze  the  audit  sample(s).   If  the 10$  criteria still
cannot  be  met,  remake the standards,  recalibrate  the GC,  and  reanalyze the audit
sample  until  the  criteria is  met or  a representative of the Administrator decides
differently.

5.3  Sample Analysis

    After the GC  has  been calibrated  and  the analysis of  the  audit sample (s) has
been conducted successfully,  the samples can  be analyzed.   Use the same procedures
for sample  analysis  that were used  to analyze the calibration standards.   Record
the GC  conditions and the analytical  data  on  the  form provided in Figure 5-1-  The
following  subsections  describe the  procedures for  analyzing  Tedlar  bag samples,
direct  and dilution interface samples,  adsorption  tube  samples, and heated syringe
samples.

5.3-1  Analysis of Bag Samples - The following procedures are to be used to analyze
emission samples collected in Tedlar  bags  using a  GC calibrated with gaseous cali-
bration  standards  prepared following  one  or more  of the  procedures  described in
Subsection 5-1-
    1.  Attach a  quick connect, or  similar  connecting device that  is compatible
        with the connection on  the Tedlar  bag to  the gas sampling valve on the GC.
        Attach a manometer connected to a tee on the outlet of  the sample loop.
    2.  With  the gas  sampling valve  in the load position,  attach  the first Tedlar
        bag sample to the valve.  Use a pump  on  the outlet side of the sample loop
        to flush the sample through the loop at 100 cc/min for  30 seconds.
    3.  Turn  off  the  pump, allow  the sample loop  to return to the same pressure
        used during calibration standard analysis,  and immediately switch the valve
        to the inject position.
    4.  Note  the  time  of  the  injection on the strip  chart  recorder and/or actuate
        the  electronic  integrator.   Also,  record  the  sample  identity,  detector
        attenuation factor, chart  speed,  sample loop temperature,  column tempera-
        ture  and  identity,  and the carrier gas type and flow  rate  on  a data form
        such  as Figure  5-1-   It is  also  recommended that  the  same information be
        recorded directly on  the chromatogram.  Record the operating parameters for
        the particular detector being used.
    5.  Examine  the  chromatogram  to ensure  that  adequate  resolution is  being
        achieved for the major components of the sample.  If adequate resolution is
        not being  achieved,  vary the  GC  conditions until  resolution is achieved,
        and reanalyze the standards to recalibrate the GC at the new conditions.
    6.  After conducting  the  analysis with  acceptable peak  resolution,  determine
        the retention time of the sample components  and compare them to the reten-

-------
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 24

        tion times for  the  standard  compounds.   To qualitatively identify an indi-
        vidual sample  component  as a  target  compound,  the retention  time for the
        component must match within 0.5 seconds or 1%, whichever is greater, of the
        retention  time  of  the  target compound  determined  with the  calibration
        standards.
     7. Repeat the  injection of  the  first  sample until  the  area count  for each
        identified target compound from two consecutive injections give area counts
        within 5 percent of their average.
     8. Multiply the average area count of the consecutive injections by the atten-
        uation factor  to get the area value  for that sample,  and record the area
        value  on  the  data  form  provided in  Figure  5-1-   NOTE: When dispensing
        liquid from  a  syringe,   take  care  to account  for the  volume  of liquid
        present in the  syringe  needle.  In general,  the  potential  error resulting
        from  the  volume  of the  needle  is most  conveniently  avoided  by ensuring
        that  the  needle volume  is  completely  full  of  liquid  upon  filling  the
        syringe and dispensing from it.   If  air  pockets  exist  in the syringe after
        filling,  this will be almost impossible.
     9- Repeat the procedure for the other two samples collected at the same sampl-
        ing location.
    10. Immediately following the analysis of  the last  sample,  reanalyze the cali-
        bration standards,  and  compare the  area values  for  each standard  to  the
        corresponding  area  values from  the  first calibration  analysis.    If  the
        individual area values  are  within 5# of their mean  value,  use  the mean
       ' values to  generate  a final  calibration  curve  for determining  the sample
        concentrations.   If the  individual values are not within 5#  of their mean
        values, generate a calibration curve using the results of the second analy-
        sis of the calibration standards,  and report the sample results compared to
        both standard curves.
    Determine  the  bag sample water  content  by measuring  the temperature  and  the
barometric pressure'near the bag.  Use water saturation vapor pressure chart, assum-
ing the relative humidity of  the  bag to  be 100% unless a  lower  value is known,  to
determine the  water vapor content as a decimal figure (% divided by  100).   If the
bag has been heated  during  sampling,  the flue gas or duct  moisture  content should
be determined using Method 4.

5-3-2   Analysis  of  Direct  Interface  Samples -  Prior  to  analysis  of  the direct
interface sample,  the  GC should  be  calibrated  using a  set  of  gaseous standards
prepared by  one  of  the  techniques  described  in  Subsection  5-1 and  a successful
analysis of an audit sample should be completed.   If possible,  the  audit samples
should be  introduced directly into  the  probe.   Otherwise,  the audit  samples  are
introduced into the  sample  line  immediately following the probe.  The calibration
is done by disconnecting the sample line  coming from the probe,  from  the gas sampl-
ing valve sample loop  inlet,  and connecting the  calibration standards  to the loop
for analysis.   During  the  analysis  of  the  calibration  standards  and  the  audit
sample(s),  make  certain  that  the sample  loop pressure  immediately  prior  to  the
injection of  the  standards  is at the same pressure  that  will be used for sample
analysis.   To  analyze the  direct interface  samples  after GC calibration,  use  the
following procedures:
     1. Reconnect the sample line to  the  inlet of the gas  sample  loop, switch the
        valve  to  the  load  position,  and turn on the sampling pump.  Adjust  the
        sampling rate to at least 100 cc/minute,  and, for  the  first  sample,  purge
        the sample  line  long enough  to  flush the sample  loop and  the preceding
        volume of tubing a minimum of 7 times.

-------
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 25

     2. After purging the sampling system  and  the  sample loop,  decrease the sample
        flow using the needle valve downstream of the loop until the loop pressure,
        measured by a water manometer connected  to  a tee at the outlet of loop, is
        equal to the pressure used during calibration.
     3- Once the  loop is at  the correct  pressure,  immediately switch  the sample
        valve to the inject position.   Note  the  time of the injection on the strip
        chart recorder and/or actuate the  electronic integrator.   The flow through
        the sample  line  can  be returned to  lOOcc/min after sample injection,  and,
        after the  unretained  compounds  are  detected,  the gas  sample  valve can be
        switched back  to the  load position.    The  system will  then be  ready to
        inject the second sample as soon as the first analysis  is completed.
     4. Record  the  sample  identity,  detector  attenuation  factor,   chart speed,
        sample loop  temperature,  column temperature  and  identity,  and the carrier
        gas type  and flow rate  on  a  form such  as  Figure 5-1-   It  is also recom-
        mended that  the  same information  be  recorded directly on the  chromatogram.
        Record the operating parameters for the particular detector being used.
     5- Examine  the  chromatogram  to   ensure  that  adequate resolution  is  being
        achieved for the major components of the sample.  If adequate  resolution is
        not being  achieved,  vary the GC  conditions  until  resolution is achieved,
        and reanalyze the standards to recalibrate the GC at the new conditions.
     6. Immediately  after the  first analysis is complete,  repeat  steps  2 and 3 to
        begin the analysis of the second sample.
     7. After  conducting the  analysis  of  the  first  sample with  acceptable  peak
        resolution,  determine   the  retention  time  of the  sample  components  and
        compare them to  the retention times  for  the standard compounds.  To quali-
        tatively identify an individual sample component as a  target  compound, the
        retention  time  for the component must  match, within  0.5 seconds  or 1%,
        whichever is greater,  the retention  time of the target compound determined
        with the calibration standards.
     8. At  the  completion of the analysis of  the second sample,  determine if the
        area counts  for  the two  consecutive injections give area  counts within 5
        percent of their average.   If this  criterion  cannot be  met due  to the
        length of  the  analysis, and the emissions  are known to vary because  of a
        cyclic or  batch  process,  then  the analysis  results  can still be used with
        the prior approval of the Administrator.
     9. Analyze a minimum of three samples collected by direct  interface to consti-
        tute an emissions test.
    10. Immediately following the analysis of  the  last sample,  reanalyze the cali-
        bration standards,  and compare the  area values  for each standard to the
        corresponding area  values  from the first  calibration  analysis.    If the
        individual  area  values  are within  5% of  their  mean  value,  use  the  mean
        values  to  generate  a  final  calibration  curve  to  determine  the sample
        concentrations.    If the individual values are not  within  5#  of their mean
        values, generate a calibration curve using the results of the  second analy-
        sis of the calibration standards,  and report the sample results compared to
        both standard curves.

5-3-3  Analysis of Dilution Interface Samples - For the analysis of dilution inter-
face samples, the procedures described  for direct  interface sampling in Subsection
5-3-2  should  be followed, with the addition  of a  check of the  dilution system.

-------
                                                              Section No. 3-16.5
                                                              Date June 30, 1988
                                                              Page 26

Prior  to  any sample  analysis,  the  GC  must first  be  calibrated, followed  by the
dilution system check and an  analysis of  the  audit  sample(s).   The audit sample(s)
are introduced  preferably  into the  inlet  to  the dilution system  or directly into
the gas sampling valve.   Use the following procedures  to  conduct the check of the
dilution system:
     1. Heat the dilution  system to the  desired temperature (0°  to  3°C above the
        source temperature) or, if  the  dilution system components can not tolerate
        that temperature, to a temperature high enough to prevent condensation.
     2. Adjust the dilution system to achieve the desired dilution rate, and intro-
        duce a high concentration target gas into the inlet of the dilution system.
        After dilution  through the  stage (s)  to  be used  for  actual  samples,  the
        target  gas  should be at a  concentration  that  is within  the calibration
        range.
     3. Purge the gas sample  loop with diluted high concentration  target  gas at a
        rate of  100  cc/min for  1 minute, adjust the  loop pressure  measured by a
        water manometer  connected to a  tee at the  outlet  of the loop,  to  the loop
        pressure that was  used during  calibration  and will be  used during sample
        analysis.  The procedure  for pressure adjustment  for the sample loop will
        vary with the type of dilution  system that is used.  In general,  the loop
        pressure can be  lowered by  reducing the  flow  into the  loop  and raised by
        restricting the flow from the loop.
     4. After achieving the proper loop pressure, immediately switch the gas sample
        valve to the inject position.
     5- Note the time of the  injection  on the strip chart recorder  and/or actuate
        the  electronic  integrator.    Also,  record  the  sample   identity,  detector
        attenuation factor, chart speed,  sample loop  temperature,  column  tempera-
        ture and identity,  and the carrier gas type and flow rate on a form such as
        Figure 5-1-  It  is  also recommended that the  same information be  recorded
        directly on  the chromatogram.    Record  the operating  parameters  for  the
        particular detector being used.
     6. Determine the peak area and retention time for the target compound used for
        the dilution check,  and calculate the area value using the detector attenu-
        ation.    Compare  the  retention  time to  the retention  time of  the target
        compound calibration standard.   The retention times should agree within 0.5
        seconds or 1%, whichever  is  greater.  If  the  retention  times do not agree,
        identify the problem and repeat the dilution check.
     7- Calculate the concentration  of  the dilution check gas  (Cd)  using  the fol-
        lowing formula.


                 Y - b
            Cd  = 	x d                                             Equation 5-1^
     where
             Y = Dilution check target compound peak area, area counts,
             b = y-intercept of the calibration curve,  area counts,
             S = Slope of the calibration curve, area counts/ppmv,  and
             d = Dilution rate of the dilution system,  dimensionless.

-------
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 27

     8. If the calculated value for the dilution check gas is not within 10% of the
        actual dilution check gas, then determine if  the  GC  or the dilution system
        is in error.  Check the calibration of the GC by analyzing one of the cali-
        bration  samples  directly  bypassing the  dilution system.    If  the GC  is
        properly calibrated, then adjust the dilution system, and repeat the analy-
        sis of  the  dilution check gas until the calculated  results  are  within 10%
        of the actual concentration.
     Once the dilution system and  the  GC  are  operating properly,  analyze the audit
sample(s).  Upon  completion of  a successful audit, the system is ready  to analyze
samples following the procedures described in Subsection 5-3-2.  To load the sample
from the  dilution system  may not require a pump on the outlet of the sample loop,
but calibration  of  the GC  using  standards  prepared in Tedlar  bags  will require a
pump.  The system should  be configured so that the pump can be taken off line when
it is not needed.

5-3-4  Analysis of  Adsorption Tube Samples - Prior  to the  analysis of adsorption
tube  samples,  the  target compounds  adsorbed  on the adsorption material  must be
desorbed.  The  procedures found to give  acceptable  desorption efficiencies deter-
mined  in  Subsection 5-1-4 should be used.  The  procedures  for the analysis of the
sample desorption solutions are  the  same as those used for  the standards.  During
sample analysis,  the  sample collection  efficiency  must be  determined.   Use the
following procedures to determine the collection efficiency:
     1. Desorb  the  primary and backup  sections  of the tubes  separately using the
        procedures  found  to  give  acceptable  (50%)  desorption efficiency  for the
        spiked adsorption  material.   Use  the  same final volume of desorption solu-
        tion for  the samples  as  was used  for  the standard solutions.  If more than
        one adsorption  tube was used  in  series  per  test run,  delay desorbing the
        additional  tubes  until  the analysis  of  the primary and  backup  section of
        the  first tube is  complete,  and the  collection  efficiency  for the first
        tube determined.   Select the  samples  from the sampling  run when the flue
        gas or  duct moisture was the  highest  and,  if known,  when  the target com-
        pound concentrations were the highest and analyze them  first.
     2. Calibrate the GC using standards prepared directly in desorption solvent or
        prepared on adsorbent and desorbed.
     3- Select  a suitably  sized injection syringe   (5-  or  10-ul),   and  flush the
        syringe with acetone  (or some other suitable solvent if acetone is a target
        compound) to clean  the syringe.
     4. Flush  the  syringe  with  the  desorption  solution from the  tube's  backup
        section by withdrawing a syringe full of the solution from the septum vial,
        and dispensing the solution into a beaker containing charcoal adsorbent.
     5- Refill  the  syringe with  the  backup section  desorption solution,  withdraw
        the syringe from  the vial, and wipe  the syringe needle  with a laboratory
        tissue.
     6. Adjust the  syringe volume down to  the amount used  for injecting standards
        and inject  the  sample into the GC.   Note the time of  the injection on the
        strip  chart recorder and/or  actuate the  electronic  integrator.   Also,
        record  the   sample identity,  detector  attenuation  factor,  chart  speed,
        injection port temperature, column temperature and identity, and the carri-
        er gas type and flow rate on the data form shown in Figure 5.1.  It is also
        recommended that the same information be recorded directly on the chromato-

-------
                                                          Section No. 3-16.5
                                                          Date June 30, 1988
                                                          Page 28

    gram. .  Record the operating  parameters  for the particular  detector being
    used.
 7- After  the  analysis,  determine  the  retention time  of  the major  sample
    components,  and compare these retention times to the retention times deter-
    mined for the target compounds during analysis of the standards.  To quali-
    tatively identify an individual  sample component as a target compound, the
    retention time  for the  component must  match, within  0.-5 seconds  or 1%,
    whichever is greater, the  retention  time of the target compound determined
    with the  calibration  standards.   Determine  the peak area for  each target
    compound identified in the sample.                            ,
 8. Repeat  the  injection  of  the  first  sample until the area counts  for each
    identified  target  compound from  two consecutive  injections are  within  5
    percent of their average.
 9. Multiply the average area count of the consecutive  injections by the atten-
    uation factor to get the area value for that sample.
10. Next analyze  the  desorption solution from the  primary  section  of the same
    adsorption tube following steps 4 through 9 above.
11. For each target compound, calculate the total weight (W),  in ug, present in
    each  section,   taking  into  account  the desorption  efficiency using the
    formula below.
         Wp or Wb  =  	 x —                                Equation 5~15
                       (Y - b)    1
                        - x —
                         S      DE                    .        .         •  .-   '

 where

         Y = Average value  for  the target compound  in  the section (primary or
             backup),  area counts,                                     • .
         b = y-intercept from the  three-point  calibration curve for the target
             compound, area counts,
         S = Slope from the three-point calibration  curve  for the target
             compound, area/ug,  and
        DE = Desorption efficiency (if standards prepared  directly in
             desorption solvent are used for calibration) .

12. Determine the  percent  of the total catch  found  in  the primary section for
    each target compound identified using the  following formula.
                    px
         E   - - x 100$                                  Equation 5-16
          cx
where
         Ecx =
               x, percent,
               Catch of co
         mbx = Catch of compound x in the backup section, ug.
         Ecx = Collection efficiency of the primary section for target compound

         mpx = Catch of compound x in the primary section, ug, and

-------
                                                              Section No.  3.16.5
                                                              Date June 30,  1988
                                                              Page 29

        If the collection  efficiency  for the primary section for  each target com-
        pound identified is >_ 90%,  then the collection efficiency for that compound
        is acceptable.   If the collection efficiency for all the  target  compounds
        identified in the sample is acceptable,  then the analysis of any additional
        tubes used in series behind the  first tube  will  not  be  necessary.   Proceed
        with the analysis of the other adsorption tube samples.
    12.  If  the  collection  efficiency  for any  identified target  compound is  not
        acceptable, then analyze the second  tube  (if  used) connected in series and
        determine the collection efficiency for  that tube using  the steps  described
        above.    If the  second  tube  does not exhibit acceptable  collection  and  a
        third tube  was  used,  analyze the  third tube.   If acceptable collection
        efficiency cannot  be demonstrated for  the sampling system,  then  the  emis-
        sion test using adsorption tubes will not be acceptable.
    13.  Immediately following the analysis of the last  sample,  reanalyze  the  cali-
        bration standards,  and compare  the  area values  for  each standard to the
        corresponding  area values from  the first  calibration   analysis.   If the
        individual area  values are within 5%  of  their mean value, use  the mean
        values to  generate a  final  calibration curve  for  determining the  sample
        concentrations.  If the individual values are not within  5# of their mean
        values,  generate a calibration curve using the results of the second analy-
        sis of the calibration standards, and report the sample  results compared to
        both standard curves.

5.3-5  Analysis of Heated  Syringe Gas  Samples by Direct Injection - For the analy-
sis of samples collected in heated syringes, the GC  will have  to  be equipped with
an injection septum fitted to  the gas  sampling  valve  sample  loop inlet.  Calibrate
the GC  following one  of  the  procedures described in  Subsection 5 • 1  for gaseous
calibration standards.   Analyze the heated syringe  samples by the following proce-
dures :
     1.  Attach a GC  septum to  a quick connect,  or  equivalent,  compatible with the
        connector on the gas sampling valve, and attach  this connector to the gas
        sampling valve.
     2.  Insert the needle  of the heated syringe through  the  septum,  and  purge the
        sample loop  by  injecting  a  volume  of  the  gas  sample  at least  ten  times
        greater than the sample loop volume.
     3.  Allow the sample loop  pressure,  measured by a water  manometer connected to
        a  tee  on the outlet of the  sample  loop, to  reach the  same loop pressure
        seen during  analysis of the calibration standards,  and immediately switch
        the gas sample valve to the inject position.
     4.  Note the time  of the injection on the  strip  chart recorder and/or actuate
        the  electronic  integrator.    Also,  record  the  sample   identity,  detector
        attenuation factor, chart speed, sample loop temperature and volume, column
        temperature and  identity, and  the  carrier gas type and  flow rate  on a form
        such as  Figure 5-l«  It is also recommended that the  same  information be
        recorded directly on the chromatogram.   Record the operating parameters for
        the particular detector being used.
     5.  Examine  the  chromatogram  to  ensure  that   adequate resolution  is  being
        achieved for the major components of the sample.  If adequate resolution is
        not being  achieved, vary the  GC conditions until resolution is  achieved,
        and reanalyze the standards to recalibrate the GC at  the new conditions.

-------
                                                          Section No. 3.16.5
                                                          Date June 30, 1988
                                                          Page 30

 6. After conducting  the analysis with  acceptable peak  resolution,  determine
    the retention time of the sample  components  and  compare them to the reten-
    tion times for the standard  compounds.   To  qualitatively identify an indi-
    vidual sample component  as  a target  compound,  the retention  time for the
    component must match, within 0.5  seconds or 1%, whichever  is  greater, the
    retention  time  of  the  target  compound  determined  with  the  calibration
    standards.
 7. Repeat the  injection of the  first sample until  the  area  counts  for each
    identified target  compound   from  two consecutive  injections are  within  5
    percent of their average.
 8. Multiply the average area count of the consecutive injections by the atten-
    uation factor to get the area value for that sample.
 9. Repeat the procedure for the other two samples  collected at the same sampl-
    ing location.
10. Immediately following the analysis of  the last sample,  reanalyze the cali-
    bration standards,  and  compare the  area values for  each  standard  to the
    corresponding area  values  from  the  first  calibration  analysis.    If the
    individual area  values  are  within 5% of  their  mean  value, use  the mean
    values to  generate  a final  calibration curve for determining  the  sample
    concentrations.   If  the  individual values  are  not within 5% of their mean
    values,  generate a calibration curve using the  results of the second analy-
    sis of the calibration standards,  and report the  sample results compared to
    both standard curves.

-------
                                                              Section No.  3.16.5
                                                              Date June 30, 1988
                                                              Page 31
Date    	 Plant Name 	'     Sampling Location 	

Checks for Analysis of All Calibration Standards

    A minimum of three concentration levels used for each target compound?
        	 yes	 no.  (The concentration used should bracket the expected
        concentrations of the actual field samples.)

   . Proper GC conditions established prior to standard analysis? 	 yes 	 no.
        (For initial  conditions use analytical  conditions found to  be acceptable
        during preliminary survey sample analysis.)

    Individual peak  areas for consecutive injections  within 5% of  their mean for
        each target -compound?  	 yes 	  no.    (Repeat analysis  of standards
        until 5% criteria is met.)

    Second analysis-, of standards after sample analysis completed? 	 yes 	 no.

    Peak areas for repeat analysis of each standard within 5% of their mean peak
       . area? 	 yes     no.  (If no, then report sample results compared to both
        standard curves.)

Checks for Calibrations using Commercial Cylinder Gases

    Vendor concentration verified by direct analysis? 	 yes 	 no.

    Sample loop purged for 30 seconds  at  100 ml/min prior to injection of calibra-
        tion standards? 	 yes 	 no.

Checks for Preparation and Use of Calibration Standards Prepared by Dilution

    Dilution system flowmeters  calibrated? 	 yes 	 no.  (Calibrate following
        procedure described in Subsection 2.1.3.)

    Sample loop purged for 30 seconds  at  100 ml/min prior to injection of calibra-
        tion standards? 	 yes 	 no.

    Dilution ratio for dilution system verified? 	 yes 	 no.   (Analysis of
        low concentration cylinder gas after establishing calibration curve
        recommended  to verify  dilution procedure,  but hot  required,  since  audit
        sample will also verify dilution ratio.)


           Figure 5-10.  Postsampling operations checklist.

-------
                                                              Section No. 3-16.5
                                                              Date June 30, 1988
                                                              Page 32
Figure 5.10 (Continued)
Checks  For  Preparation  and  Use of  Calibration Standards  by Direct  Injection of
Gaseous Compounds or Liquid Injection

    Tedlar bag used to contain prepared standard leak and contamination free?
        yes 	 no 	.

    Dry gas meter used to  fill  bag calibrated?  	 yes 	 no.   (Calibrate meter
        following procedure described in Subsection 2.1.2.)

    Organic standard material used for injection 99-9$ pure? 	 yes 	no.  (If
        no, then determine purity and use to correct calculated calibration
        standard concentration.)

    Prepared standard allowed to equilibrate prior to injection? 	 yes 	 no.
        (Massage bag by alternately depressing opposite ends 50 times.)

    Sample loop purged for 30 seconds  at  100  ml/min prior to injection of calibra-
        tion standards? 	 yes 	 no.


Development of Relative Response Factors and Retention Times

    Suitable target organic or surrogate compound selected? 	 yes 	 no.
        (Select compound that  is stable, easy  to  prepare in the field,  and has  a
        retention time similar to the target organic compounds.)

    Relative response factors and retention times  verified in the laboratory prior
        to actual field use? 	 yes 	 no.  (If no, verify following the
        procedure described in Subsection 5-1-4-)


Checks for Preparation,  Use,  and Determination of Desorption Efficiency for Adsorp-
tion Tube Standards

    Organic standard material used for injection 99-9% pure? 	 yes 	 no.  (If
        no,  then determine purity and use to correct calculated calibration
        standard concentration.)

    Correct adsorbent material  and desorption solvent selected? 	  yes 	 no.
        (Refer to Table B in Method Highlights Section for proper adsorbent
        material and desorption solvent.)

    Desorption efficiency determined for adsorbent to be used for field sampling?
        	 yes 	 no.  (If no,  follow the  procedure described in Subsection
        5.1.5.)

-------
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 33
Figure 5.10  (Continued)

Checks'for All GC Analysis of Field Samples

    Check type of carrier gas used: helium 	,  nitrogen 	, other
    Carrier gas flow rate and pressure set correctly?  	 yes 	 no.  (Carrier
        gas flow rate and pressure set according to conditions developed during
        .presurvey/  sample  analysis and  within limitations  of the GC  as specified
        by GC manufacturer.)

    Oxygen and hydrogen flow rate and pressure for FID correct? •_.	 yes _.	 no
      .  (Oxygen and-hydrogen gas flow.rate and pressure for FID set according to
        conditions developed during presurvey sample analysis and within
        limitations of the GC as specified by GC manufacturer.)

    Individual peak  areas.. for consecutive  injections  within 5% of  their mean for
        each  target  compound? 	  yes  	 no.   (Repeat  analysis  of standards
        until 5?» criteria is met.)   ,        -...-        .     •  •,          \

    Audit sample analyzed and results within 10% of actual value? 	 yes 	 no.
        (If no, recalibrate GC and/or reanalyze audit sample.)

Checks Type of Standard Used for Tedlar Bag Sample Analysis
                        ;                                       '
    Gas cylinders 	._, dilution of gas cylinders	, direct.gas injection 	,
        direct liquid injection• _.	, and/or relative response factors and
        retention times 	.

Checks.For GC Analysis Of Tedlar Bag Samples                  •  •       '

    Sample loop purged for  30  sec.  at 100 ml/min prior to injection of calibration
        standards? 	 yes 	 no.

    Stability of gas  sample in Tedlar bag determined? 	 yes-      no. -..-. (Deter-
        mine  stability by conducting a second  analysis  after the first  at a time
        period equal  to  the,time between collection and the  first  analysis.   The
        change in  concentration- between,  the first  and  second analysis  should be
        less than 10%.)                                •          ,   .  :

    Retention of- target compounds in  Tedlar  bag determined?      yes	 no.  (If
       •no, then follow.the procedure described in Subsection 5.3.1.)

Check GC Interface Technique Used
                   J              i " '        *        : '             '   t        " ' '
    Direct Interface .•,  10:1 Dilution Interface   , 100:1 Dilution Interface

-------
                                                               Section No.  3.16.5
                                                               Date  June  30,  1988*
                                                               Page  34

 Figure 5-10  (Continued)

. Checks For Suitability ,of GC  Interface Technique

     Analytical interference due  to moisture content of source gas?  	  yes  	 no.
          (Moisture  in  the  source  gas  must not interfere with analysis in regard
        to peak resolution according  to EPA Method 625 criterion where the
          baseline-to-valley height between adjacent peaks is less  than 25$  of the
          sum  of the two adjacent  peaks.)

     Physical requirements for equipment  met on-site?  	 yes 	 no.   (The
          physical requirements for the equipment include sheltered environment,
          "clean", uninterrupted power source  suited for equipment,  and adherence  to
          safety aspects related to explosion risk areas.)

     Source gas concentration  below level of  GC detector saturation? 	yes  	 no.
          (Concentrations delivered  to the detector can be  reduced by using  smaller
          gas  sample loops and/or  dilution interface.)

     Sampling systems  purged with 7 changes of  system volume prior to sample
          analysis?  	 yes 	no.

 Check Type(s) of Standards Used  for  Interface  Techniques

     Gas  cylinders  	, dilution of  gas  cylinders 	, direct gas  injection 	,
          direct liquid injection  	, and/or relative response factors and
          retention  times 	.

 Checks For Dilution Interface Analytical Apparatus

     Dilution rate  verified  (within 10%) by introducing  high  concentration gas
          through dilution system  and  analyzing  diluted gas? 	 yes 	 no.
          (If  dilution rate  not  verified,  then  first  check  calibration  of GC  by
          reanalyzing a calibration  standard and then adjust dilution system  to give
          desired ratio).

     Sampling systems  purged with 7 changes of  system volume prior to sample
          analysis?  	 yes 	no.

 Check Type of Standard Used for  Adsorption Tube Analysis

     Prepared directly in  desorption  solvent  	,  and/or prepared  on adsorbent and
          desorbed 	.

 Checks for GC Analysis of Adsorption Tube Samples

     Desorption procedure used identical  to procedure used to determine the
          desorption efficiency? 	  yes  	 no.

-------
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 35
Figure 5-10  (Continued)
    Collection efficiency  determined for  adsorption tubes  used for  actual field
        sampling? 	 yes  	 no.   (If no,  then  determine collection efficiency
        following the procedures described in Subsection 5-3-4.)

Check Type of Standard Used for Analysis of Heated Syringe Samples

    Gas cylinders 	,  dilution of gas cylinders 	,  direct gas injection 	,
        direct liquid injection 	, and/or relative response factors and
        retention times

-------
                                                               Section  No.  3.16.5
                                                               Date  June  30,  1988
                                                               Page  36
                  Table 5.1.  ACTIVITY MATRIX FOR SAMPLE  ANALYSIS.
Characteristic
 Acceptance limits
 Frequency  and  method
    of  measurement
 Action..if  . .
 requirements
 are not met
Calibration
Standards
All calibrations
1) Standard analysis
performed under, same
GC conditions to be
used for,samples

2) Three-point
(minimum), calibra-
tion .curve generated
for each .target
compound

3) .Sufficient amount
of each standard to'
recalibrate after
samples are analyzed
Before  analysis  of
calibration  standards
determine .samplei.ana- .
lysis conditions

Before  analysis  Ac-
quire or prepare stan-
dards for each target
•compound at  three, •
levels

Prior to initial • ,
calibration  and  sample
analysis, determine-
amount  needed
 Reanalyze stan-
 dards under con-
ditions to be
 used for samples

 Acquire or pre-
 pare standards at
 at  three levels
 to  bracket
 samples
                                                                  Acquire..,or
                                                                  prepare
                                                                  enough :
                                                                  standards
Commercial gas
cylinder mixtures
Certified by .direct
analysis (within 5$
of manufacturer,' s
value); three levels
bracketing samples
Prior ,to use,, check  if
''independent analysis
conducted and accept-
able and standards,
will bracket  samples
 Procure certified
 gas  cylinders in
 proper range
Gas standards from
high concentration
gas cylinders
Dilution ratio, of  .:
dilution system     i
verified .(optional)
with calculated val-
ue using calibration
curve within 10%
of actual cone,.
Prior  to  sample  analy-
sis ; calibration curve
from standards verif-
ied by analysis  of  an
undiluted sample •
• Identify, and
 correct problems
 with dilution
 system,  and
 remake,  reana-
 lyze, and re-
 verify, standards
Standards prepared
by direct gas
injection
Gas injected 99-9%
pure, or calculated
standard concentra-
tion corrected for
gas impurity
When  calculating  stan-
dard  concentration,
determine purity  of
gas standard      >
 Use  pure gas or
 determine purity
Standards prepared
by liquid
injection  ,
(Continued)
Liquid injected
99-9$ pure, or.
calculated standard
corrected 'for
liquid impurity
•When calculating.stan-
dard concentration,
determine purity of
liquid standard.
 Use  pure liquid
 or determine
 purity

-------
Table 5.1  (Continued)
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 37
Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Calibration
Standards

Relative response
factors and rela-
tive retention
times
Proper target or
surrogate standard
selected for on-site
calibration; method
verified (calculated
results within 10%
of actual concentra-
tion)
When selecting stan-
dard choose stable,
easy to prepare stan-
dard with retention
time near or between
target compounds; ver-
fied following proced-
ures described in
Subsection 5-1-5
Select different
target or
surrogate
compound; if
procedure cannot
be verified
use calibration
standard for each
target compound
Standards prepared
for adsorption
tube samples'
1) Liquid injected
99.92 pure, or
calculated standard
corrected for

2) Acceptable
desorption effici-
ency for target com-
pounds on adsorbent
material (>50#)
When calculating stan-
dard concentration,
determine purity of
liquid standard

During calibration
standard analysis
determine desorption
efficiency for each
target compound (see
Subsection 5.1.5)
Use pure liquid
or determine
purity
                                                                  Try longer de-
                                                                  sorption times,
                                                                  more vigorous
                                                                  desorption condi-
                                                                  tions , and/or
                                                                  other desorbents
Audit sample
analysis
Analytical result
for audit sample
within 10% of actual
concentration
After initial cali-
bration and prior to
sample analysis, ana-
lyze audit sample
Reanalyze audit
sample, if not
acceptable, re-
make and reana-
lyze standards
Sample Analysis

All samples
1) Audit sample
analysis within
of actual cone.
                    2) Sample analysis
                    conditions the same
                    as conditions used
                    for analysis of
                    standards
Prior to sample ana-
lysis, analyze audit
sample

Prior to sample ana-
lysis check that ana-
lytical conditions are
the same as those used
for standard analysis
Analyze audit
sample
                                              Establish the
                                              same analytical
                                              conditions used
                                              during analysis
                                              of standards
(Continued)

-------
Table 5.1  (Continued)
                                                              Section No. 3-16.5.
                                                              Date June 30, 1988
                                                              Page 38
Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Sample Analysis

All samples
3) Retention times
for target compounds
identified in sample
within 0.5 seconds
or 1% of standards
                    4) Area counts for
                    consecutive injec-
                    tions of samples
                    within 5% of their
                    average for each
                    target compound
                    identified in sample
                    5) All three samples
                    constituting a test
                    analyzed together

                    6) After sample ana-
                    lysis , repeat analy-
                    sis of standards;
                    area counts for each
                    standard analysis
                    within 5# of their
                    mean
After analysis, deter-
mine retention times
for major components
in sample and compare
to standard retention
times

After second analysis
of a sample, calculate
average area for first
and second analysis
and percent difference
of single analysis
from the average
                      During sample analysis
                      After analysis of last
                      sample repeat standard
                      analysis; calculate
                      mean area counts and
                      percent difference for
                      each standard
Qualitative
identification
requires reten-
times within 0.5
seconds or 1%;
repeat analysis

Repeat sample
injections until
consecutive in-
jections are
achieved meeting
the 5% criteria
for each target
compound

Analyze remaining
samples
                        Report sample
                        results using
                        both curves, if
                        5# criteria not
                        met
Bag samples
1) Bag sample moist-
ure content deter-
mined
                    2) Stability check
                    conducted on bag
                    content (
-------
Table 5.1  (Continued)
                                                              Section No. 3.16.5
                                                              Date June 30, 1988
                                                              Page 39
Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Sample Analysis

Direct interface
samples.
Two consecutive
injections give
area counts within
5# of their mean
After second analysis,
calculate average area
counts and percent
difference
Due to cyclic or
batch processes
and analysis
time, emission
levels may vary;
use results with
the prior appro-
val of the Ad-
ministrator
Dilution interface
samples
1) Dilution ratio
verified (results
from analysis of
high concentration
standard through
dilution system
within 10% of actual
concentration

2) Two consecutive
injections give
area counts within
5# of their mean
Prior to sample ana-
lysis analyze high
concentration gas
introduced through
dilution system
                                          After second analysis,
                                          calculate average area
                                          counts and percent
                                          difference
Identify problem;
recalibrate GC or
adjust dilution
system and repeat
analysis of high
concentration
gas
                        Due to cyclic or
                        batch processes
                        and analysis
                        time, emission
                        levels may vary;
                        use results with
                        the prior ap-
                        proval of the
                        Administrator
Adsorption tube
samples
Collection effici-
ency determined for
adsorption tubes
OQ% of each target
compound identified
caught on primary
section)
Desorb and analyze
primary and backup
sections separately
Analyze addi-
tional tube(s) if
used as backups
to first tube; if
criteria cannot
be met, test is
not valid

-------
                                                                  Section No.  3.16.6
                                                                  Date June  30,  1988
                                                                  Page 1
6.0    CALCULATIONS
      Calculation errors due to procedural  or mathematical  mistakes  can  be  a part  of
 total system error.   Therefore, it is recommended that each set  of calculations  be
 repeated or spotchecked, preferably  by  a team member other  than the one  who  per-
 formed the original calculations.   If a difference greater than typical  round-off
 error is detected,  the calculations should be checked  step-by-step  until  the source
 of error is found  and corrected.   A computer program is advantageous in  reducing
 calculation errors.   If a standardized computer program is used, the original  data
 entered should be included  in  the  printout so it can be reviewed;  if differences
 are observed,  a new computer  run  should  be made.   Table 6.1 at  the  end  of  this
 section summarizes  the quality assurance activities  for calculations.
      Calculations should be carried out  to at least  one extra  decimal figure beyond
 that of the acquired data and  should be rounded off after final  calculation to two
 significant digits  for each  run  or  sample.   All  rounding of  numbers  should  be
 performed in accordance with  the ASTM 380-76 procedures.   All calculations should
 then be recorded on a calculation form such as the ones in Figures  6.1 and 6.2 for
 analysis by gas  or  liquid injection,  respectively.

 6.1  Calculations for GC Analysis Using  Gas Injection

      The same  equation can  be  used to calculate  the concentration  of  each organic
 in Method 18 samples  whenever  the  sampling technique used yields  a gaseous sample
 which can be injected  into  the GC.   These techniques are:  (1)  the integrated bag
 sampling technique,   (2)  the heated bag sampling  technique,  (3)  the prefilled bag
 sampling technique,  (4) the direct interface sampling technique, and (5)  the dilu-
 tion interface sampling technique.    This  equation is used to calculate the sample
 concentration  (Cc)  in ppm on a dry  basis as follows:

                 Cs  Pr  T,  Fr  K
        Cc   =  	                                        Equation 6-1
               Pi  Tr  (1 - Bws/K)
 where
      Cs   =  Concentration of organic from calibration  curve,  ppm,
      Pr   ~  Reference pressure,  the barometric  pressure  or absolute  sample  loop
             pressure recorded during calibration,  mm Hg,
      Ti   =  Sample loop temperature at time of  sample  analysis,  °K,
      Pi   =  Barometric or absolute sample loop  pressure  at time  of sample analysis,
             mm Hg,
      Tr   =  Reference temperature, the temperature of  the  sample loop  recorded
             during calibration,  °K,
      Bws  =  Water vapor content  of the stack gas,  proportion  by  volume,
      Fr   =  Relative response factor,  if applicable (see Subsection  5.1.5),  and
      K   =  Dilution factor (applicable only for dilution  interface  and  prefilled
             bag sampling;  for a  10 to  1 dilution,  K =  10).

-------
                                                                 Section No. 3.16.6
                                                                 Date June 30, 1^88
                                                                 Page 2
6.2  Calculations for GC Analysis by Liquid Injection
     For Method  18,  liquid injection GC analyses are used  in  conjunction with the
adsorption tube  sampling procedure.   The same general  equations are typically used
to calculate the concentration of  each  organic in a sample  collected on an adsorp-
tion tube.   However, the  tester is  referred to  the National  Institute of Occupa-
tional  Health  and  Safety  (NIOSH)  method (see  Table  B  in the  Method Highlights
Section) for specifics on  calculations  for particular  organics.  The general equa-
tions are shown below.

6.2.1  Sample Volume Corrected to Standard Conditions on a Dry Basis - The correct-
ed sample volume (V8td idry) is calculated as shown.

                     T    P    V                      P    V
                      std rbar vm                ^    rbar vm
     vs        =  -  =  0.3858 -
                   Pstd Ts  (1 - Bws/K)'              Ts  (1 -  BWS/K)'
                                                                       Equation 6-2

where
     Tstd/pstd  =  °-3858 °K/mm Hg,
     Vm         =  Sample volume measured, L,
     Pbar       =  Barometric pressure during sampling, mm Hg,
     Ts         =  Temperature of sample gas, °K,
     Bws        =  Water vapor of stack gas, proportion by volume, and
     K          =  Dilution factor, if applicable.

'Note:  Only apply this correction if a dessicant is not used.


6.2.2  Desorption Efficiency  - Desorption efficiency  (DE)  for recovery of a speci-
fic compound  using  a certain solvent  from an adsorption  tube  is calculated using
the following equation.


                   Q  - B
                    r
          DE  =  -                                             Equation 6-3
where
     Qr  =  Average peak area for spiked tubes,
     Qa  =  Average peak area for spiked solutions, and
     B   =  Average peak area for media blanks.


6.3.3  Concentration of Organic in Sample - The concentration (C) of the organic in
the sample in milligrams per dry  standard  cubic meter or micrograms per dry stand-
ard liter (mg/dscm or ug/dsL) is calculated using the following equation.

-------
                                                                 Section No. 3.16.6
                                                                 Date June 30, 1988
                                                                 Page 3

            (W, + Wb - Bf - BJK
     C  =  	                                      Equation 6-4
                 Vstd * DE


where                        -                                                     .

     W '   •     =  Mass of organic found in primary sorbent section, ug,
     Wb        =  Mass of organic found in backup sorbent section, ug,
     B         =  Mass of organic found in primary section of average media
                  blank, ug,
     Bb        =  Mass of organic found in backup section of average media
                  blank, ug,
     K         =  Dilution factor, if applicable (for a 10 to 1 dilution, K = 10),
     vstd dry  =  Sample volume corrected to standard conditions and a dry
                  basis, L, and
     DE        =  Desorption efficiency, decimal value.


6.2.4   Conversion  to  ppm  -  To  convert  the concentration  in milligrams  per dry
standard  cubic meter  (micrograms per  dry  standard liter)  to ppm,  the following
equation can be used.


      r     -  24.055 (dsL/g-mole gas) x C                                  .   ,
      C     =  	    	  	 	                             Equation 6-5
                          MW
where
     C   =  Concentration of organic, ug/dsL or mg/dscm, and
     MW  =  Molecular weight of organic, ug/ug-mole.

-------
              •  '<"                                                Section  No.  3.16.6
                                                                  Date  June  30,  1§88
                                                                  Page  4

                                SAMPLE CONCENTRATION

Cs  =	l_ _£_ 0_  ppm,    Pr   =   _?• JT3^ .  SL  mm Hg,   T.  = . 2>_ _7  0_ . _O  °K,



K*  =  A/ _A . _,   F/   =  fi/A_ _
               Cs Pr T. Fr  K
       Cc  =  -  =  __ _7^ _4 ppm                       Equation 6-1
              pi Tr  (1  - Bws/K)*
*If applicable.
          Figure 6.1.  Calculation  form  for GC analysis by gas injection.

-------
                                                                  Section No. 3.16.6
                                                                  Date June 30,  1988
                                                                  Page 5
                  SAMPLE  VOLUME,  DRY BASIS AT STANDARD CONDITIONS

Vm  =  __ 2-5^ .  2- L,    Pbar   =  J7 J[~+_ •  j[ mm Eg,
                          P    V
V         =  0.3858  - — — - -  =     2-^.0   L             Equation 6-2
 s t d i d r y                             _       -  ^~^~ ^—   _^_
                       Ts  (1  -  Bws/K)*

'If applicable.
                                DESORPTION EFFICIENCY

Qr  =  _# "L .1  •    Qa   =  -1 _T_^ ,    B  =	^



     DE  =   (Qr  -  B)/Qa   =  0 .  _f_ _^        .                           Equation 6-3



                                 SAMPLE  CONCENTRATION

Wp  =  _6 1_ _k  .  0_ ug,    Wb  =  _/_ _£ ^ . _^ ug,   Bp  =	^_ ug,

B.  =            ^ ug,    V ..  =      Z^.^L,  DE  =  0  .  ^ 2-  .   .
         (Wp  + Wb  -  Bp  - Bb)K
 C  =  - =  __ ^ J2_ •  _*L mg/dscm or ug/dsL        Equation 6-H
                                  CONVERSION TO PPM

 C  =   ___ .  _  mg/dscm or ug/dsL,   MW  =  __ ;£ _^  . _/  ug/ug-mole,


 _       24.033  (dsL/g-mole gas)   x C          / -Z_    -?               _    t .    £  _
 Cppm  = - =i^J - La - s -   =  __ LTL •  ±Z  PPm          Equation 6-5
                  MW


         Figure 6.2.  Calculation  form  for GC  analysis  by  liquid injection.

-------
                                                                 Section No.  3.16.6
                                                                 Date June 30,  1988
                                                                 Page 6
                Table 6.1.  ACTIVITY MATRIX FOR CALCULATION CHECKS
Characteristic
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Analysis data
form
All data and calcu-
tions are shown
Visually check
Complete the
missing data
Calculations
Difference between
check and original
calculations should
not exceed round-off
error
Repeat all calcula-
tions starting with
raw data for hand
calculations; check
all raw data input
for computer calcu-
lations; hand calcu-
late one sample per
test
Indicate errors
on calculation
form, Figure 6.1
or 6.2

-------
                                                                 Section No. 3.16.7
                                                                 Date June 30,  1988
                                                                 Page 1


7.0     MAINTENANCE

      The normal use of  emission-testing  equipment subjects it to corrosive gases,
 extremes in  temperature,  vibration,  and shock.   Keeping the equipment  in  good
 operating order over an extended period of time requires knowledge of the equipment
 and a program of  routine  maintenance  which  is performed quarterly or after 2830  L
 (100 ft3) of operation, whichever  is  greater.   In addition to the quarterly main-
 tenance, a yearly cleaning of  pumps  and metering systems  is  recommended.   Main-
 tenance procedures  for  the  various components are summarized  in  Table 7.1 at  the
 end of the section.   The following procedures are not required, but are recommended
 to increase the reliability of the equipment.

 7.1  Pump

      Several  types of pumps may be used  to perform Method 18; the two most common
 are the fiber vane pump with in-line oiler and  the diaphragm pump.  The fiber  vane
 pump requires a periodic  check  of the oiler jar.  Its contents should be translu-
 cent; the oil should be changed if not translucent.   Use the oil specified by  the
 manufacturer.   If none  is  specified, use  SAE-10 nondetergent oil.  Whenever a fiber
 vane pump  starts to run  erratically   or  during the  yearly disassembly,  the  head
 should be removed and  the fiber vanes changed.   Erratic  operation of a diaphragm
 pump is normally due to either a bad diaphragm (causing  leakage) or to malfunctions
 of the valves,  which should be cleaned annually by complete disassembly.

 7.2  Dry Gas  Meter

      Dry gas  meters  should be  checked  for excess oil  or  corrosion of the components
 by removing the  top plate every 3  months.   Meters  should be  disassembled and  all
 components cleaned  and checked whenever the  rotation of the dials  is  erratic,
 whenever the meter  will not calibrate  properly over the required flow rate range,
 and during the  yearly maintenance.

 7-3  Rotameter

      Rotameters should  be disassembled and cleaned according to the manufacturer's
 instructions  using only recommended cleaning fluids every 3 months or upon erratic
 operation.

 7 A  Manometer

      The fluid  in the manometers should be changed whenever there is discoloration
 or visible matter in the fluid,  and during the yearly disassembly.

 7-5  Sampling Train

      All remaining  sampling  train  components  should be. visually  checked  every  3
 months  and completely  disassembled and  cleaned  or replaced yearly.   Many items,
 such as quick disconnects, should be replaced whenever damaged rather than checked
 periodically.   Normally, the best procedure  for maintenance in  the field is to have
 on hand another entire  unit such as a pump,  Tedlar bags and containers,  or heated
 sample line rather than replacing individual  components.

-------
                                                                 Section No. 3.14-7
                                                                 Date June 30, 1988
                                                                 Page 2
7.6  Gas Chromatograph
     Maintenance activities and schedules for gas chromatographs are make and model
specific.   It  is  therefore recommended  that the  analyst consult  the  operator's
manual for instructions relative to maintenance practices and procedures.

-------
                                                    Section No.  3-16.7
                                                    Date  June  30,  1988
                                                    Page  3
Table 7.1.  ACTIVITY MATRIX FOR EQUIPMENT MAINTENANCE CHECKS

Apparatus
Fiber vane
pump
Diaphragm
pump
Dry gas meter
Rotameter
Manometer
Sampling
train
components
Gas chroma-
tograph

Acceptance limits
In-line oiler
free of leaks
Leak- free valves
functioning properly
No excess oil,
corrosion, or er-
ratic rotation of
the dial
Clean and no erra-
tic behavior
No discoloration or
visible matter in
the fluid
No damage
See owner ' s manual

Frequency and method
of measurement
Periodically check
oiler jar; remove
head and change fiber
vanes
Clean valves during
yearly disassembly
Check every 3 mo. for
excess oil or corro-
sion by removing the
top plate; check
valves and diaphragm
yearly and whenever
meter dial runs erra-
tically or whenever
meter will not cal-
ibrate
Clean every 3 mo. or
whenever ball does
not move freely
Check periodically
and during disassemb-
ly
Visually check every
3 mo.; completely
disassemble and
clean or replace
yearly
See owner ' s manual

Action if require-
ments are not met
Replace as
needed
Replace when
leaking or mal-
functioning
Replace parts as
needed, or replace
meter
Replace
Replace parts
as needed
If failure noted,
replace appro-
priate components
See owner ' s manual

-------
                                                              Section No. 3.16.8
                                                              Date June 30, 1988
                                                              Page 1
8.0     AUDITING  PROCEDURES
     An  audit  is  an  independent  assessment  of  data  quality.    Independence  is
 achieved  if  the  individual(s)   performing  the  audit  and  their  standards  and
 equipment  are  different  from the  regular  field  team  and  their  standards  and
 equipment.    Routine quality  assurance checks  by a field  team  are necessary to
 generate good quality data,  but they are not part  of  the auditing  procedure.  Table
 8.1 at  the end  of  this  section  summarizes  the  quality assurance  functions for
 auditing.
     Based on the requirements  of  Method 18 and the results of collaborative test-
 ing of other Reference Methods, two  specific performance audits are recommended:
         1.  An audit of the sampling and analysis of Method 18 is  required  for NSPS
            and recommended for other purposes.
         2.  And audit of the data processing is  recommended.
     It is suggested  that  a  systems  audit  be conducted  as specified  by  the quality
 assurance coordinator in addition  to these performance  audits.  The two  performance
 audits and  the  systems audit are described in  detail  in  Subsections 8.1  and 8.2,
 respectively.

 8.1  Performance Audits

     Performance  audits are  conducted  to  evaluate  quantitatively the  quality of
 data produced by the total measurement system  (sample  collection,  sample  analysis,
 and data processing).    It  is required  that  cylinder gas performance audits be
 performed once  during  every NSPS test utilizing  Method 18 and  it  is  recommended
 that  a  cylinder gas  audit  be performed  once during  any  enforcement  source  test
 utilizing Method 18 conducted under  regulations other than NSPS.

 8.1.1  Performance Audit of the Field Test - As stated  in Section  6.5 of 40 CFR 60,
 Appendix A, Method  18, immediately  after  the preparation of the calibration curves
 and prior to the sample analysis,  the  analysis audit described  in  40  CFR 61,  Appen-
 dix C,  Procedure 2:  "Procedure  for Field Auditing GC Analysis,"  should be per-
 formed.   The information  required to  document the analysis of the audit  sample(s)
 has been included  on  the  example data sheets  shown in Figures  8.1 and  8.2; the
 complete text of the procedure is reproduced  in Section 3•16.10.   The  audit anal-
 yses shall  agree within  10  percent  (or other  specified value,  as  explained  below)
 of  the   true value.    When  available,  the tester  may obtain audit cylinders by
 contacting: U.S. Environmental Protection  Agency,  Atmospheric Research and Exposure
 Assessment  Laboratory,  Quality Assurance Division  (MD-77B), Research  Triangle Park,
 North Carolina 27711•  Audit cylinders obtained from a commercial  gas manufacturer
 may be used provided that (1)  the gas manufacturer certifies the  audit  cylinder in
 a manner similar to  the  procedure described in 40 CFR 61,  Appendix  B,  Method 106,
 Section  5-2.3-1.  and  (2)  the gas  manufacturer obtains an  independent  analysis.
 Independent analysis  is  defined   as an analysis performed by  an  individual other
 than  the individual  who performs  the gas manufacturer's  analysis,  while using
 calibration standards and analysis equipment different from those used  for the gas
 manufacturer's  analysis.    Verification   is  completed  and  acceptable  when  the
 independent analysis concentration  is  within  5 percent of  the gas  manufacturer's
 concentration.

     Responsibilities of  the Audit  Supervisor  - The primary  responsibilities of
 the audit supervisor are  to ensure  that  the  proper  audit  gas  cylinder(s) are or-

-------
                                                              Section No. 3.16.8,,
                                                              Date June 30, 1988
                                                              Page 2

dered and safe-guarded, and to interpret the results obtained by the analyst.
    When auditing sampling systems  that  do  not  dilute  the stack gases during samp-
ling, the audit gases ordered must consist of the same organic compound(s) that are
being  tested;  for  emission  standards  on  a concentration  basis,  the  audit  gas
concentration(s)  must be in the range of 25% to 250% of the applicable standard. If
two  cylinders  are not  available,   then  one cylinder  can be used.   If  the audit
cylinder value is between  5  and 20 ppm, the agreement should be within 15 percent
of the stated audit cylinder value.  It is strongly recommended that audit cylinder
values below  5  PPm not be  used.    For  emission standards which specify  a control
efficiency, the concentration of  the  audit gases should  be  in  the range  of 25% to
250% of the expected stack gas  concentration.   If ttoo  cylinders are not available,
the audit can be conducted using one cylinder.
    The audit supervisor must ensure  that .the audit gas cylinder(s) are shipped to
the  correct  address,  and to  prevent  vandalism, verify  that they are  stored  in a
safe location both  before .and after  the audit.  Also,  the  audit  cylinders should
not be analyzed when  the  pressure drops below  200  psi.   The audit supervisor then
ensures that the audits are conducted as described below.
    The audit supervisor must also  interpret the audit results.   When the measured
concentration agrees within  10  percent (or 15  percent  for cylinders between 5 and
20 ppm)  of the true value,  he  directs  the  analyst to begin analyzing the source
samples.   When  the measured  concentration does  not  agree within  the  specified
criterion, the analyst should first recheck the analytical system and calculations,
and  then repeat  the  audit.   If  the analyst  fails  the  second audit,  the audit
supervisor  should have  knowledge  of the  agency's policy  for failure.    If  the
result(s) are close to  the allowed percentage or a consistent bias is present, the
supervisor may wish to allow the analyst use of a correction  factor to be applied
at a later date;  however, the  analyst  must make a  significant  effort  to find the
discrepancy  and  correct it.   If.the error  cannot be  found, the  audit supervisor
should allow analysis of the samples, and then  conduct the audit again.
    During the audit,  the audit supervisor should  record the  appropriate cylinder
number(s), cylinder pressure(s)  (at the  end of  the. audit),  and the calculated con-
centrations  on  the "Field audit  report form",   Figure 8.1.   The  individual being
audited must not, under any  circumstances,  be  told the actual audit concentrations
until the  calculated  concentration(s) have been submitted to  the  audit supervisor
and are considered acceptable.
  ... -When  auditing sampling  systems that dilute the  emissions  during collection,
the-audit, gas concentration  value used in the  calculations  can either be based on
(!•)'  the undiluted  concentration using  the criteria  discussed above  or  (2)  the
expected concentration  of  the gases following dilution during collection using the
same dilution factor as used for the emission samples.
    The audit procedures  that follow are presented  according to the type of samp-
ling  system  used  to  collect  the  organic  emissions  and  whether  the  samples  are
analyzed on-site or at the base laboratory at a.later date.

    Container (Bag,  Syringe, and  Canister) Sampling  with  On-site Analysis  -  The
cylinder gas performance  audit  for rigid-container bag, syringe, or canister samp-
ling with on-site analysis consists of  an  on-site audit just prior to the analysis
of the emission  samples.   The recommended procedures  for conducting the audit are
as follows:
        1..  The audit  samples should  be collected  in  the  type of  container that
            will be used during the sample collection.  However, to conserve on the
            use of  the  audit gas(es), it is usually not  necessary to use the rest
            of the  sampling system to  collect  the  samples  for unheated container

-------
                                                         Section No. 3.16.8
                                                         Date June 30, 1988
                                                         Page 3
                            FIELD AUDIT REPORT

Part A. - To be filled out by organization supplying audit cylinders.
          1. Organization supplying audit sample (s) and shipping address
4.
5-
6.
                                                    j
          2. Audit supervisor', organization, and phone number
             Shipping instructions: Name, Address, Attention
                           100 Pn>ke Ave,
             Guaranteed arrival date for cylinders -
             Planned shipping date for cylinders -
             Details on audit cylinders from last analysis

a. Date of last analysis 	
b. Cylinder number 	
c . Cylinder pressure psi 	
d. Audit gas (es) /balance gas..
e. Audit gas(es) , ppm 	
f . Cylinder construction 	

Low cone.
Iz-Jitofoft
IQb'k
[S'LrO
.£•+&, /Jf/2~
A4

High cone



.C.+.h'/^i
Zi'so
A&

Part B. - To be filled out by audit supervisor.
          1. Process sampled
          2. Audit location     $,fe.
          3. Name of individual audit
          4. Audit date 	
          5. Audit Results:

a. Cylinder number 	
b. Cylinder pressure before audit, psi 	
c. Cylinder pressure after audit, psi 	
d. Measured concentration, ppm
Injection #1* Injection #2* Average 	
e. Actual audit concentration, ppm
f. Audit accuracy:1
Low Cone. Cylinder 	
High Cone. Cylinder 	
Percent1 accuracy =
Measured Cone. - Actual Cone. x 100
Actual Cone.
g. Problems detected (if any) 	

Low
cone.
cylinder
/Od>^
/57>£>
/S7?0
#l/*#>
+S.$%


hM

High
cone.
cylinder
/%&
p^IDD
2^?*®
2lx>/V4v

£>?*>

n^

          1 Results of two consecutive injections that meet the sample analysis
           criteria of the test method.
                  Figure 8.1.  Field audit report form.

-------
                                                              Section No.  3.16.8
                                                              Date June  30,  1988
                                                              Page 4

            sampling.  Problems related to the reaction or retention of  the  organic
            compounds will still occur in the container.  Other interferents in the
            stack gas such as water vapor and other organics will not be present in
            the  audit  cylinders and  thus,  related problems will  not be  assessed.
            For  heated  container  systems,  it may  be  necessary  to use the sampling
  •   v       system to collect  the  audit  gas.   However,  if the gases must-be heated
            to prevent  condensation,  it  is likely that  an  audit gas cylinder will
            not  be available.
        2.  The  audit  samples should  remain  in the  appropriate  container  approx-
            imately  the same  length  of  time  that  the  source samples will  stay
            prior to analysis.   After the preparation of the calibration curve,  a
            minimum of  two  consecutive analyses of each  audit  cylinder gas should
            be conducted.   The analyses  must agree within 5% of  the average.   The
            audit results should  be calculated by the  analyst  (or representative)
            and  given  to the audit  supervisor.   The audit  supervisor  will record
            all  the information and data on the"Field audit report form" and then
            inform  the   analyst  of the  status of  the  audit.   The  equations for
            calculation  of error are included on the form.

    Container (Bag and Canister) Sampling with Off-site Analysis - For cylinder gas
performance  audits  associated  with rigid-container  bag  or canister  samples  that
are analyzed off-site,  it is recommended that  the audit be conducted off-site just
prior  to  the emission  test  (if the  agency  desires)  and then  repeated during the
off-site  sample  analysis as a quality  control  measure.   The  use of   the  pretest
audit  will help ensure that  the  analytical  system will  be acceptable  prior to
testing.   Alternatively, the  audit gas  can be  collected in the  appropriate con-
tainer  on-site  or off-site,  and  then analyzed just  prior to  the  analysis  of the
field samples.   It is recommended that the tester fill at least two containers with
the audit gas to guard against a leak causing a failed audit.  Since the use of the
performance audit is  to both assess  and  improve  the data quality,  the  use  of the
pretest audit  will  provide  the tester/analyst with  a  better  chance  of obtaining
acceptable data.  The recommended procedure for conducting the audit is  the  same as
above  with the  exception  that  the audit  supervisor will  likely not  be  present
during the audit and the data will be reported by telephone.

    Direct Interface Sampling - Since direct  interface  sampling  involves  on-site
analysis,  the performance audit is conducted on-site after  the calibration  of the
GC and prior to  sampling.   The audit gas cylinder is attached  to  the inlet  of the
sampling probe.   Two consecutive  analyses  of  the audit  gas must be within 5% of
the average of  the  two  analyses.   The tester/analyst then  calculates   the  results
and informs the audit supervisor.   The audit supervisor records all information and
results on the "Field audit  report  form"  and then informs the tester/analyst as to
the acceptability of the results.

    Dilution Interface  Sampling -  Since  dilution interface sampling  involves on-
site analysis,  the performance  audit  is  conducted on-site after the calibration of
the GC and prior to sampling.   If  the audit  gas  cylinder obtained has a concentra-
tion '-near  the  diluted  sample concentration, the  audit  gas  is  introduced directly
into the sample  port on the  GC.   If the audit gas cylinder  obtained  has a concen-
tration close to the  expected sample  concentration,  then the audit  gas is  intro-
duced into the dilution  system.  The  audit supervisor may wish  to order one  cylin-
der to assess both the  dilution system and the analytical system and another cylin-
der to  assess  only the  analytical system.  Follow  the  same procedures described

-------
                                                              Section No. 3.16.8
                                                              Date June 30, 1988
                                                              Page 5

above for recording the information and reporting the results.

    Adsorption Tube Sampling - The analysis for adsorption tube sampling is usually
conducted off-site.   Therefore,  the audit analysis is conducted  off-site.   Again,
the recommended procedure is to conduct  the audit  once prior  to the test and again
following  the  test.    Though  the  audit  sample  could be  analyzed  by  direct
injection,  the inclusion of the chromatogram printout in  the report will prove that
the  audit  results were  obtained  through  adsorption tube  sampling and  a solvent
extraction.  Alternatively, the audit  samples  can  be collected on-site or off-site
and then analyzed just prior to the analysis of the field samples.  Since the audit
supervisor will likely not be present during the analysis, the results are reported
by telephone.
    To collect  the audit gas  with the  adsorption tube  sampling  train,  connect a
sample "T"  to the line  from the  audit  gas cylinder.   Place the  adsorption tube
sampling system on one leg of  the  "T"; connect a rotameter to the other leg.  With
the  sampling system  off, turn on  the audit gas  flow until the  rotameter reads 2
1pm. Turn  on the sampling system  and sample  the  audit  gas for  the specified run
time.  Approximately 1 1pm should be discharged through the rotameter.

8.1.2  Performance Audit of Data Processing -  Calculation errors are prevalent in
processing data.  Data processing errors can be determined by auditing the recorded
data on  the  field "and laboratory  forms.   The  original  and audit (check) calcula-
tions should agree within round-off error; if not, all of the remaining data should
be  checked.   The data  processing may  also  be audited  by providing  the testing
laboratory with specific data sets (exactly as would appear  in  the field), and by
requesting that the data calculation  be  completed  and that the results be returned
to  the agency.  This  audit  is useful  in  checking both computer programs and manual
methods of data processing.

8.2  Systems Audit

    A systems audit  is an on-site, qualitative inspection and review of the total
measurement  system  (sample  collection,   sample   analysis,  etc.).    Initially,  a
systems  audit  is  recommended for  each enforcement source test,  defined here as a
series of  three runs  at  one  source.   After the test team gains experience with the
method, the  frequency of auditing may be reduced — for example,  to once every four
tests.
    The  auditor should  have  extensive  background experience in source sampling,
specifically  with the  measurement system being  audited.    The  functions  of the
auditor are  summarized below:
     1.  Inform  the  testing team of the  results of pretest audits,  specifying any
        area(s)  that need special attention or  improvement.
    2.  Observe procedures and techniques of the  field  team  during sample collec-
        tion.
    3-  Check/verify  records  of apparatus  calibration  checks and  quality control
        used  in the  laboratory  analysis of control  samples   from  previous source
        tests, where applicable.
    4.  Record  the  results of  the audit,  and forward  them  with comments  to the
        test  team  management  so that  appropriate corrective  action  may  be
        initiated.
    While  on site, the  auditor observes  the   source  test team's  overall perfor-
mance, including the following specific operations:

-------
                                                              Section No. 3-16.8
                                                              Date June 30, 1988
                                                              Page 6

    1.  Conducting the GC calibration and  conducting the performance audit (if the
        analysis is conducted on-site).
    2.  Setting up and leak testing the sampling train.
    3.  Collecting the  sample at a proportional  rate (if applicable)  or constant
        rate at the specified flow rate.
    4.  Conducting the final leak check and recovery of the samples.
    5.  Conducting the  initial  and final  check on the dilution  system (if appli-
        cable) .
    6.  Sample  documentation procedures,  sample  recovery,  and  preparation  of
        samples for shipment  (if applicable).
    7.  Conducting sample analyses (if conducted on-site).
Figure 8.2 is a suggested checklist for the auditor.

-------
                                                            Section No. 3-16.8
                                                            Date June 30, 1988
                                                            Page 7
Yes
No
Comments
Operation
               A//4
                                 PRESAMPLING PREPARATION
                     1.  Knowledge of process operations
                     2.  Results of pretest audit (_+ 10% or other value)
                     3.  Calibration of pertinent equipment, in
                         particular, dry gas meters and other flowmeters
                     4.  Selection and checkout of equipment for proper
                         sampling and analytical techniques
                         BAGS - reactivity, condensation, & retention
                         ADSORPTION TUBES - adsorption & desorption
                                            efficiency
                         DILUTION SYSTEM - dilution ratio
                         GC/COLUMN - adequate resolution
                         GC/DETECTOR - acceptable accuracy & precision
               V/4
              N/A
              V/A
                                  ON-SITE MEASUREMENTS
                     5.  Results of on-site audit (*_ 10% or other value)
                     6.  Sampling system properly assembled
                     7.  Based on pitot tube check, is proportional
                         sampling required (more than 10% flow change)
                     8.  Dilution system check acceptable (if applicable)
                     9.  Sampling system leak check acceptable
                    10.  Proportional sampling properly conducted
                    11.  Constant rate sampling properly conducted
                    12.  Heater systems maintained at proper temperatures
                    13.  Proper number of samples & sampling time
                    14.  GC properly calibrated
                    15.  Duplicate injections had acceptable precision <5%
                    16.  Recording of pertinent process conditions during
                         sample collection, samples properly identified,
                         and calculations properly conducted
               rilA
               V/A
              A//*
                                     POSTSAMPLING
                    17.  Results of off-site audit (+_ 10% or other value)
                    18.  GC properly calibrated
                    19.  Duplicate injections had acceptable precision <5%
                    20.  Adsorption efficiency acceptable,>90% on primary
                    21.  Desorption efficiency acceptable,>50% recovery
                    22.  Adequate peak resolution
                    23.  Bags passed reaction check,  less than 10% change
                    24.  Bags passed retention check,less than 5% retained
                    25.  Flowmeters recalibration acceptable
                    26.  Temperature sensor recalibration acceptable
                                              COMMENTS
                                                a-

        Figure 8.2.  Method 18 checklist to be used by auditors.

-------
                                                              Section No. 3.16.8
                                                              Date June 30, 1988
                                                              Page 8
             Table 8.1.  ACTIVITY  MATRIX  FOR AUDITING PROCEDURES
Apparatus
 Acceptance limits
Frequency and method
   of measurement
Action if
requirements
are not met
Performance
audit of
analytical phase
Measured relative
error of audit
samples less than
10% (or other stated
value) for both
samples
Frequency;  Once during
every enforcement
source test*
Method: Measure audit
samples and compare
results to true values
Review operating
technique and
repeat audit
Data processing
errors
Original and checked
calculations agree
within round-off
error
Frequency:  Once during
every enforcement
source test*
Method: Independent
calculations starting
with recorded data
Check and correct
all data for the
audit period
represented by
the sampled data
Systems audit—
observance
of technique
Operational tech-
nique as described
in this section of
the Handbook
Frequency;  Once during
every enforcement
source test* until
experience gained,
then every fourth
test
Method; Observation of
techniques assisted
by audit checklist,
Figure 8.1
Explain to team
their deviations
from recommended
techniques and
note on Fig 8.1
*As defined here,  a source test for enforcement of the NSPS comprises a series of
 runs at one source.  Source test for purposes other than enforcement of NSPS may
 be audited at the frequency determined by the applicable group.

-------
                                                               Section No. 3-16.9
                                                               Date June 30, 1988
                                                               Page 1
9.0     RECOMMENDED STANDARDS  FOR  ESTABLISHING  TRACEABILITY

      To  achieve  data  of  desired  quality,   two  essential  considerations  are
 necessary:     (1) the measurement process must bu in a state of  statistical control
 at the time  of the  measurement, and  (2)  the systematic  errors, when combined with
 the  random  variation   (errors  or  measurment),   must  result  in an  acceptable
 uncertainty.   As evidence  in support of  good quality  data,  it  is  necessary to
 perform  qulaity  control  checks  and  independent audits of the measurement process;
 to  document  these data;  and  to  use  materials,  instruments,   and  measurement
 procedures that can be traced to an  apropriate standard of reference.

      Data  must be  routinely obtained by repeat measurements of standar reference
 samples   (primary,  secondary, and/or working standards)  and the establishment of a
 condition of  process  control.    The  working calibration  standards  should  be
 traceable  to standards of higher accuracy.

      Audit samples  (as  discussed  in Section 3-16.8) must be used  to validate test
 results  for  compliance  determination  purposes   and   are  recommendeed  as  an
 independent  check on the measurement process when the method is  performed for other
 purposes.

-------
                                                                                      Section No.  3-16.10
                                                                                      Date June 30,  1988
                                                                                      Page 1
10.0    REFERENCE   METHOD'
        Since   the  initial  promulgation  of  Method  18  in  1983   (48  FR,  48344   -  48360,
 10/18/83),   there  have  been  a number of  revisions  and  additions  to  the method.    In
 the  interest  of  consistency and  clarity,  the  version  of  Method  18  reproduced here
       from  the  most   recent  edition   of   the  Code   of  Federal  Regulations   which
 is
 incorporates  all promulgated changes to this date.
METHOD 18—MEASUREMENT OF GASEOUS OR-
  GANIC COMPOUND EMISSIONS BV GAS Ciiso-
  MATOGRAPHY

Introduction
  This method should not be attempted by
persons  unfamiliar with the performance
characteristics of gas chromatography. nor
by those persons who  are unfamiliar with
source sampling.  Particular care should be
exercised In the  area of safety concerr.'r.g
choice of equipment and operation In poten-
tially explosive atmospheres.
1. Applicability and Principle
  1.1  Applicability. This method applies to
the analysis of approximately 90 percent of
the total gaseous organics emitted from an
Industrial source. It does not Include tech-
niques  to  identify  and  measure  trace
amounts  of  organic compounds, such as
those found In  building air  and  fugitive
emission sources.
  This  method  will not  determine  com-
 pounds that (1) are polymeric (high molecu-
 lar weight). (2) can polymerize before  analy-
 sis, or (3) have very low vapor pressures at
 stack or Instrument conditions.
   1.2 Principle.
  The major organic components of  a gas
mixture are separated by gas chromatogra-
phy  (GO and individually quantified by
flame ionization, photoionization,  electron
capture,  or  other  appropriate detection
 principles.
  The  retention times of each separated
 component  are  compared with those  of
 known  compounds  under identical  condi-
 tions. Therefore, the  analyst  confirms the
 identity and approximate concentrations of
 the  organic emission  components before-
 hand. With  this information, the analyst
 then prepares  or purchases  commercially
 available standard mixtures to calibrate the
 GC under conditions  identical to  those of
 the samples. The analyst also determines
 the need for sample dilution to avoid detec-
 tor saturation, gas stream filtration to elimi-
 nate particulate matter, and prevention of
 moisture condensation.
 2. Range and Sensitivity
   2.1  Range. The range of this method is
 from about 1 part per million (ppm) to the
 upper limit governed by GC detector satura-
 tion or column overloading. The upper limit
 can be extended by diluting the stack gases
 with an inert gas or by using smaller gas
 sampling loops.
  2.2 Sensitivity. The sensitivity limit for a
compound Is defined  as the minimum de-
tectable concentration of that compound, or
the concentration that produces a signal-to-
nolse ratio of three to one.  The minimum
detectable  concentration   Is  determined
during  the  presurvey calibration for each
compound.
3. Precision and Accuracy
  Gas chromatographlc techniques typically
provide a precision of 5 to 10 percent rela-
tive standard deviation (RSD). but an expe-
rienced GC operator with a  reliable Instru-
ment can readily achieve 5 percent RSD.
For this  method, the following combined
GC/operator values are required.
  (a)  Precision.  Duplicate  analyses  are
within  5 percent of their mean value.
  (b) Accuracy. Analysis results of prepared
audit samples are within 10 percent of prep-
aration values.
4. Interferences
  Resolution Interferences that may occur
can be  eliminated  by  appropriate  GC
column and detector choice or by shifting
the retention times through changes in the
column flow rate and the use of tempera-
ture programming.
  The  analytical system Is demonstrated to
be essentially free from contaminants by pe-
riodically analyzing  blanks  that consist of
hydrocarbon-free air or nitrogen.
  Sample cross-contamination that  occurs
when  high-level and low-level samples  or
standards are analyzed alternately,  is best
dealt with by thorough purging of the GC
sample loop between samples.
  To assure consistent detector response.
calibration  gases are contained In dry air.
To adjust  gaseous organic  concentrations
when water vapor Is present In the sample,
water  vapor concentrations  are determined
for those samples, and a correction factor Is
applied.
5. Presurvey and Presurvey Sampling.
  Perform a presurvey for each source to be
tested. Refer to Figure 18-1.  Some of the  In-
formation can be collected  from literature
surveys and source  personnel.  Collect gas
samples that can be analyzed to confirm the
identities and approximate  concentrations
of the  organic emissions.
  S.I   Apparatus. This apparatus list also
applies to Sections 6 and 7.
 *40  CFR 60,  Appendix A,  Method  18,  July  1,  198?,  pages  ?40  - 769.

-------
                                                                                              Section No.  3.16.10
                                                                                              Date  June  30,  1988
                                                                                              Page  2
  5.1.1  Teflon Tubing.  (Mention of trade
names  or specific products does not consti-
tute endorsement by the U.S. Environmen-
tal   Protection  Agency.)  Diameter  and
length  determined  by connection require-
ments  of cylinder regulators and the  GC.
Additional  tubing  is necessary to connect
the GC sample loop to the sample.
  S.I.2  Gas Chromatograph. GC with suit-
able detector,  columns,  temperature-con-
trolled sample loop and valve assembly, and
temperature programable oven, if necessary.
The GC shall achieve  sensitivity require-
ments for the compounds under study.
  5.1.3  Pump. Capable of pumping 100 ml/
mln. For flushing sample loop.
  5.1.4  Flowmeters. To measure flow rates.
  5.1.5  Regulators. Used  on  gas  cylinders
for GC and for cylinder standards.
  5.1.6  Recorder. Recorder with linear strip
chart  Is minimum acceptable. Integrator
(optional) is recommended.
  5.1.7  Syringes. 0.5-ml,  1.0- and  10-mlcro-
llter sizes,  calibrated, maximum  accuracy
(gas tight), for preparing calibration stand-
ards. Other appropriate sizes can be used.
  5.1.8  Tubing Fittings. To plumb GC and
gas cylinders.
  5.1.9  Septums. For syringe injections.
  5.1.10 Glass Jars. If necessary,  clean-col-
ored glass  jars  with  Teflon-lined lids for
condensat.3 sample collection. Size depends
on volume of condensate.
  5.1.11 Soap Film Flow Meter. To deter-
mine flow rates.
  5.1.12 Tedlar Bags. 10- and 50-liter capac-
ity, for preparation of standards.
  5.1.13 Dry Gas Meter with Temperature
and Pressure Gauges. Accurate to ±2 per-
cent, for perparatlon of gas standards.
  5.1.14 Midget  Implnger/Hot  Plate As-
sembly. For preparation of gas standards.
  5.1.15 Sample Flasks. For presurvey sam-
ples, must have gas-tight seals.
  5.1.16 Adsorption Tubes.  If  necessary,
blank tubes filled with necessary adsorbent
(charcoal. Tenax, XAD-2, etc.) for presur-
vey samples.
  5.1.17 Personnel  Sampling  Pump. Cali-
brated, for collecting adsorbent tube presur-
vey samples.
  5.1.18 Dilution  System. Calibrated, the
dilution system Is to be constructed  follow-
ing  the specifications  of  an  acceptable
method.
  5.1.19 Sample Probes. Pyrex or stainless
steel, of sufficient length to reach centroid
of stack, or a point no closer to the walls
than 1 m.
  5.1.20 Barometer. To measure barometric
pressure.
  5.2 Reagents.
  5.2.1  Delonlzed Distilled Water.
  5.2.2  Methylene DIchloride.
  5.2.3  Calibration Gases. A series of stand-
ards prepared for every compound of inter-
est.
  5.2.4-'Organic Compound Solutions. Pure
(99.9 percent), or as pure as can reasonably
be obtained, liquid samples of all the organ-
ic compounds needed to prepare calibration
standards.
   5.2.5 Extraction Solvents. For extraction
 of adsorbent tube samples in preparation
 for analysis.
   5.2.6 Fuel. As recommended by the man-
 ufacturer for operation of the GC.
   5.2.7 Carrier Gas. Hydrocarbon free,  as
 recommended by the manufacturer for op-
 eration of  the  detector  and compatabllity
 with the column.
   5.2.8 Zero Gas. Hydrocarbon free air  or
 nitrogen, to be used  for  dilutions, blank
 preparation, and standard preparation.
   5.3  Sampling.
   5.3.1 Collection of Samples with Glass
 Sampling Flasks.  Presurvey samples can  be
 collected  In precleaned  250-ml  double-ended
 glass  sampling  flasks.  Teflon  stopcocks,
 without grease,  are preferred. Flasks should
 be cleaned as follows: Remove the stopcocks
 from both ends of the  flasks, and wipe the
 parts to remove any grease. Clean the stop-
 cocks, barrels, and receivers with methylene
 dlchloride. Clean all glass ports with a soap
 solution, then rinse with tap and delonlzed
 distilled water.  Place the  flask in a  cool
 glass annealing  furnace and apply heat up
 to 500* C. Maintain at this  temperature for
 1  hour. After this time period, shut off and
 open the  furnace  to allow the  flask to cool.
 Grease the stopcocks with  stopcock grease
 and return  them to the  flask receivers.
 Purge the assembly with high-purity nitro-
 gen for 2 to 5 minutes. Close  off the stop-
 cocks  after  purging  to maintain  a  slight
 positive nitrogen pressure. Secure the stop-
 cocks with tape.
   Presurvey samples can be obtained either
 by drawing  the gases  Into  the previously
 evacuated flask or by drawing the gases Into
 and purging the flask with a rubber suctloi.
 bulb.
  5.3.1.1  Evacuated Flask Procedure. Use a
 high-vacuum pump to evacuate the flask to
 the capacity of the pump; then  close off the
 stopcock leading to the pump. Attach a 6-
 mm outside diameter (OD) glass tee to the
 flask  Inlet  with a short piece of Teflon
 tubing. Select a  6-mm OD borosilicate sam-
 pling probe, enlarged at one  end to a 12-mm
 OD and of sufficient length to reach  the
 centroid of the duct to be sampled.  Insert a
 glass wool plug  in the enlarged end of the
 probe to remove particulate matter. Attach
 the other end of the probe to the tee with a
 short  piece  of  Teflon  tubing. Connect  a
 rubber suction bulb to the third leg of the
 tee. Place  the filter end  of the probe at the
 centroid of the duct, or  at a point no closer
 to the walls than 1 m. and purge the probe
 with  the  rubber  suction bulb. After  the
 probe is completely purged  and filled with
 duct gases, open the  stopcock  to the grab
 flask until the pressure in the flask reaches
duct pressure. Close off the stopcock, and
 remove the probe from  the duct. Remove
 the tee from the  flask  and tape the stop-
cocks  to  prevent  leaks  during shipment.
 Measure and record  the duct temperature
and pressure.
  5.3.1.2  Purged  Flask  Procedure.  Attach
one end of the sampling flask  to a rubber
suction bulb. Attach the other end to a 6-
mm OD glass probe as described in  Section

-------
                                                                                            Section  No.  3-16.10
                                                                                            Date  June  30,  1988
                                                                                            Page  3
5.3.1.1. Place the filter end of the probe at
the centroid of the duct, or at a point no
closer to the walls than 1 m. and apply suc-
tion with the bulb to completely purge the
probe and  flask. After the flask  has been
purged, close off the stopcock near the suc-
tion bulb, and then close the stopcock near
the probe. Remove the  probe from the duct,
and disconnect both the probe and suction
bulb. Tape  the stopcocks to prevent leakage
during shipment. Measure  and record the
duct temperature and pressure.
  5.3.2  Flexible Bag Procedure. Tedlar or
aluminlzed  Mylar bags can also be used to
obtain the  presurvey sample. Use new bags,
and leak check them before field use. In ad-
dition, check the bag  before use  for  con-
tamination by filling it with nitrogen or air,
and analyzing the gas by GC at high sensi-
tivity. Experience indicates that it Is desira-
ble to  allow the Inert gas to remain  In the
bag about  24 hours or longer to check for
desorptlon  of organics from the bag. Follow
the leak check and  sample collection proce-
dures given in Section 7.1.
  5.3.3  Determination  of Moisture Content.
For combustion or water-controlled process-
es, obtain the moisture content from plant
personnel  or  by  measurement during the
presurvey.  If  the  source Is below  59' C,
measure the wet bulb and dry bulb tempera-
tures,  and  calculate the moisture content
using a psychrometric chart. At higher tern
peratures,  use Method 4  to determine the
moisture content.
   5.4   Determination   of  Static  Pressure.
Obtain the static pressure  from the plant
personnel or measurement. If a type S pilot
tube  and an Inclined  manometer are used,
take  care  to align  the pltot tube 90*  from
the direction of the flow. Disconnect one of
the tubes  to the manometer, and read the
static pressure: note whether the reading  is
positive or negative.
   5.5  Collection of Presurvey Samples with
Adsorption Tube. Follow Section 7.4 for pre-
survey sampling.
 6. Analysis Development
   6.1  Selection of GC Parameters.
   6.1.1 Column Choice. Based on the initial
 contact with plant  personnel concerning the
 plant process and the  anticipated emissions,
 choose a column that provides good resolu-
 tion and rapid analysis time. The choice  of
 an appropriate column can be aided by a lit-
 erature search, contact with manufacturers
 of GC columns, and discussion with person-
 nel at the  emission source.
   Most column manufacturers  keep excel-
 lent records of their products. Their techni-
 cal service departments may be able to rec-
 ommend appropriate  columns and detector
 type  for  separating  the  anticipated com-
 pounds, and they may be able to provide in-
 formation on interferences, optimum oper-
 ating conditions, and column limitations.
   Plants with analytical  laboratories may
 also  be able  to provide information on ap-
 propriate  analytical procedures.
   6.1.2  Preliminary GC Adjustment. Using
 the standards and column obtained in Sec-
 tion 6.1.1, perform initial tests to determine
 appropriate  GC  conditions that  provide
 good resolution and minimum analysis time
 for the compounds of interest.
  6.1.3  Preparation of Presurvey Samples.
 If the samples were collected on an adsorb-
 ent, extract the sample as recommended by
 the manufacturer for removal of the com-
 pounds with a solvent suitable to the  type
 of GC analysis. Prepare other samples In an
 appropriate manner.
  6.1.4  Presurvey Sample Analysis. Before
 analysis, heat the presurvey sample to the
 duct temperature to vaporize any condensed
 material.  Analyze the samples by the  GC
 procedure, and compare the retention times
 against those  of  the calibration samples
 that contain the components expected to be
 in the stream. If any compounds.; cannot be
 identified with certainty by this procedure,
 identify them by other means such as GC/
 mass spectroscopy (GC/MS) or GC/lnfrared
 techniques. A GC/MS  system  is recom-
 mended.
  Use the GC conditions determined by the
 procedures of Section 6.1.2 for the first in-
 jection. Vary the  GC parameters during
 subsequent injections to determine the opti-
 mum settings.  Once  the optimum settings
 have been determined, perform repeat injec-
 tions of the sample to determine  the reten-
 tion time  of each compound.  To inject a
 sample, draw sample through  the  loop at a
 constant rate (100 ml/min for 30 seconds).
 Be careful not to  pressurize the gas In  the
 loop. Turn off the pump and allow the gas
 In the sample loop to come to ambient pres-
 sure. Activate the sample valve, and record
 Injection  time,  loop temperature, column
 temperature, carrier flow rate, chart speed.
 and attenuator setting. Calculate  the reten-
 tion time of each peak using the distance
 from injection to the peak maximum divid-
 ed  by the chart speed. Retention times
 should be repeatable within 0.5 seconds.
  If the concentrations are too high for ap-
 propriate   detector   response,  a  smaller
 sample loop or dilutions may be used for gas
 samples, and. for liquid samples, dilution
 with solvent is appropriate.  Use the stand-
 ard  curves (Section  6.3) to  obtain an  esti-
 mate of the concentrations.
  Identify  all  peaks  by  comparing  the
 known retention times of compounds  ex-
 pected to be in the retention times of peaks
 in the sample.  Identify any remaining un-
 identified  peaks which have  areas  larger
 than 5 percent of the total using a GC/MS.
 or  estimation of  possible compounds  by
 their retention times compared to known
 compounds, with confirmation by further
 GC analysis.
  6.2 Calibration  Standards.  Prepare  or
 obtain enough calibration standards so that
 there are  three different concentrations of
 each organic  compound  expected to  be
 measured in the source sample. For each or-
 ganic compound, select those concentrations
 that bracket the concentrations expected In
 the source samples. A calibration standard
may contain more than one organic com-
pound.  If  available,  commercial  cylinder
 gases may  be  used If their  concentrations
have been certified by direct analysis.
  If samples are collected in adsorbent tubes
(charcoal,  XAD-2,  Tenax, etc.). prepare or
obtain standards in the same  solvent used
for the sample extraction procedure. Refer
to Section 7.4.3.

-------
                                                                                                Section  No.  3.16.10
                                                                                                Date  June  30,  1988
                                                                                                Page  4
    Verify the stability  of all standards  for
  the time periods they are used. If gas stand-
  ards are prepared in the laboratory, use one
  or more of the following procedures.
    6.2.1  Preparation of  Standards  from
  High Concentration  Cylinder  Standards.
  Obtain enough high concentration cylinder
  standards to represent  ail the organic com-
  pounds expected in the. source samples.
    Use these  high concentration standards to
  prepare lower  concentration standards by
  dilution, as shown by Figures 18-5 and 18-6.
    To  prepare  the  diluted calibration sam-
  ples.  calibrated rotameters  are normally
  used to  meter both the high concentration
  calibration gas and the diluent  gas. Other
  types of flowmeters and commercially avail-
  able dilution systems can also be used.
    Calibrate each flowmeter  before  use  by
  placing  it between  the  diluent gas supply
  and suitably sized bubble meter, spirometer.
  or wet test meter. Record all data shown on
  Figure 18-4. While it is desirable to calibrate
  the cylinder  gas flowmeter with cylinder
  gas, the available  quantity  and cost may
  preclude it. The error introduced by using
  the diluent gas for calibration is insignifi-
  cant for  gas mixtures of up to 1,000 to 2,000
  ppm of each organic component.
   Once the flowmeters  are calibrated, con-
  nect the flowmeters to  the calibration and
  diluent  gas  supplies using  6-mm  Teflon
  tubing. Connect the outlet side of the flow-
  meters through a connector  to  a leak-free
  Tedlar bag as  shown in Figure  18-5. (See
  Section 7.1 for bag leak-check procedures.)
  Adjust the gas flow to  provide the desired
  dilution,  and fill the bag with sufficient gas
  for GC calibration. Be careful not to overfill
 and cause the bag to apply additional pres-
 sure on the dilution system. Record the flow
 rates of both flowmeters, and the laborato-
 ry temperature  and atmospheric pressure.
 Calculate the concentration  C. in ppm of
 each organic in the diluted gas as follows:
106  (X  q
                             )
             Eq .    18-1
 where:
 10" = Conversion to ppm.
 X=Mole or volume  fraction of the organic
    In the calibration gas to be diluted.
 qc=Flow rate of the  calibration gas to be di-
    luted.
 qd= Diluent gas flow  rate.
 Single-stage dilutions should be used to pre-
 pare calibration mixtures up to about 1:20
 dilution factor.
  For greater dilutions, a double dilution
 system is recommended, as shown in Figure
 18-6. Fill the Tedlar  bag with the dilute gas
 from the second stage.  Record the laborato-
 ry temperature,  barometric  pressure, and
static pressure readings. Correct  the  flow
reading for temperature and pressure. Cal-
culate the  concentration C, in ppm  of the
organic  in the final gas mixture as follows:
                                                                                               Eq.    18-2
 Where:
 10'=Conversion to ppm.
 X=Mole or volume  fraction of the organic
    in the calibration gas to be diluted.
 q,i=Flow rate of the calibration gas to be di-
    luted In stage 1.
 qd=Flow rate of the calibration gas to be di-
    luted in stage 2.
 q«i=Flow rate of diluent gas In stage 1.
 q,n=Flow rate of diluent gas in stage 2.
  Further details of the calibration methods
 for flowmeters and the dilution system can
 be found In Citation 21 in the Bibliography.
  6.2.2  Preparation   of  Standards  from
 Volatile Materials. Record all data shown on
 Figure 18-3.
  6.2.2.1  Gas  Injection  Technique. This
 procedure  is  applicable  to organic com-
 pounds that exist entirely as a  gas at ambi-
 ent conditions. Evacuate  a 10-liter Tedlar
 bag that has passed a leak-check (see Sec-
 tion 7.1), and meter in 5.0 liters of air or ni-
 trogen through a dry  gas meter that  has
 been calibrated in a manner consistent with
 the procedure described In Section 5.1.1 of
 Method 5. While the bag is filling use a 0.5-
 ml  syringe to inject  a known quantity of
 "pure"  gas  of  the   organic   compound
 through the  wall of the bag. or through a
 septum-capped tee at  the  bag inlet. With-
 draw  the syringe needle, and  Immediately
 cover the  resulting  hole with  a piece of
 masking tape. In a like manner, prepare di-
 lutions  having other  concentrations. Pre-
 pare a minimum of  three  concentrations.
 Place each  bag on a smooth surface, and al-
 ternately depress opposite sides of the  bag
 50 times to mix the gases Record the aver-
age meter  temperature and pressure,  the
 gas volume and  the  barometric  pressure.
Record the syringe temperature and pres-
sure before Injection.
  Calculate each  organic  standard concen-
 tration C. in pom as follows:
                                                                               r     in6 293
                                                                               by  x 10   	
                               P

                               760
         Vm  Y    291
                                                                                Gv x  ID"
                                                                                              1000
                                                                                                            Eq.   18-3

-------
                                                                                              Section No.  3.16.10
                                                                                              Date  June  30,  1988
                                                                                              Page  5
 where:
 O,=Gas volume or organic compound  In-
    jected, ml.
 ,o*= Conversion to ppm.
 P.=Absolute pressure of syringe before  in-
    jection, mm Hg.
 T.=Absolute temperature of syringe before
    Injection. 'K.
 Va-Gas volume Indicated by dry gas meter.
    liters.
 Y=Dry gas meter calibration factor, dlmen-
    slonless.
 ?„=Absolute pressure of dry gas meter, mm
    Hg.
 Tm=Absolute temperature of dry gas meter,
    •K.
 1000=Conversion factor, ml/liter. .
  6.2.2.2  Liquid Injection  Technique. Use
 the equipment shown in Figure  18-8.  Cali-
 brate the dry gas meter as described in Sec-
 tion 6.2.2.1 with a wet test  meter or a spi-
 rometer. Use  a water manometer for the
 pressure gauge and glass. Teflon, brass, ot
 stainless steel for all connections. Connect a
 valve to  the inlet of the 50-liter Tedlar bag.
  To  prepare the standards,  assemble the
 equipment as  shown In  Figure 18-8, and
 leak-check the system. Completely evacuate
 the bag. Fill the bag with hydrocarbon-free
 air, and  evacuate the bag again. Close the
 Inlet valve.
  Turn on  the hot plate,  and  allow the
 water to reach boiling.  Connect  the bag to
 the  Implnger  outlet.   Record  the initial
 meter reading, open the bag inlet valve, and
 open  the cylinder.  Adjust the rate, so that
 the bag will be completely filled in approxi-
 mately IS minutes.'Record meter pressure
 and temperature, and local barometric  pres-
 sure.
  Allow the liquid organic to  equilibrate to
 room temperature. Fill the 1.0- or 10-micro-
 llter syringe to the desired liquid volume
 with'the organic. Place the syringe needle
 Into the  impinger  inlet using the septum
 provided, and inject the liquid into the  flow-
 Ing air stream. Use a needle of sufficient
 length  to permit  Injection  of  the. liquid
 below the  air inlet branch  of the tee.
 Remove the syringe.
  When the bag is filled, stop the pump, and
 close the bag inlet valve. Record the  final
 meter reading, temperature, and pressure.
  Disconnect  the  bag  from  the impinger
 outlet, and either set It aside for at least 1
 hour, or massage the bag to Insure complete
 mixing.
  Measure  the  solvent liquid density at
room temperature by accurately weighing a
known volume of the material on an analyt-
ical balance  to  the nearest 1.0 milligram.  A
ground-glass  stoppered  25-mil  volumetric
flask  or  a glass-stoppered  specific  gravity
bottle Is suitable for weighing. Calculate the
result in  terms of g/ml. As  an alternative,
literature values of the density of the liquid
at 20 *C may be used.
  Calculate each organic standard concen-
tration C. in  ppm as  follows:
(24.055 x  10)
                         6.24 x  10
293   P
                                             Y  P
     750
            100°
                                       Eq.   18-4
       where:
       L.= Liquid volume of organic injected. pi.
       p.l= Liquid organic density as determined, g/
          ml.
       M= Molecular weight of organic, g/g-mole.
       24.055 = Ideal gas molar volume  at 293 *K
          and 760 mm Hg, llters/g-mole.
       10'= Con version to ppm.
       1000= Con version factor,
        6.3  Preparation  of Calibration Curves.
       Establish proper GC conditions, then flush
       the sampling loop for 30 seconds at a rate of
       100 ml/min. Allow the sample loop pressure
       to  equilibrate to atmospheric pressure, and
       activate the Injection valve.  Record  the
       standard concentration, attenuator factor,
       Injection time, chart speed, retention time,
       peak area, sample loop temperature, column
       temperature, and  carrier  gas flow  rate.
       Repeat the standard Injection until two con-
       secutive Injections give area counts within 5
       percent of their average. The average value
       multlpled by the attenuator factor is then
       the calibration area value for the concentra-
       tion.
        Repeat this procedure for each  standard.
       Prepare a graphical  plot of concentration
       (C.) versus the calibration area values. Per-
       form a  regression  analysis, and  draw  the
       least squares line.
        6.4  Relative Response Factors.  The cali-
       bration curve generated from the standards
       for a single organic can usually be related to
       each  of the  individual GC  response curves
       that are developed in the laboratory for all
       the compounds in the source.  In  the field,
       standards for that single organic  can then
       be used to "calibrate" the GC for all the or-
       ganics present. This procedure should first
       be confirmed in the laboratory by preparing
       and analyzing calibration standards contain-
       ing multiple organic compounds.
        6.5  Quality  Assurance  for  Laboratory
       Procedures. Immediately after  the prepara-
       tion of the  calibration curves and prior to
       the presurvey sample analysis, the analysis
       audit described in 40 CFR Part 61, Appen-
       dix C, Procedure  2: "Procedure for  Field Au-
       diting GC Analysis," should be performed.
       The information  required to document  the
       analysis of the audit samples has been  in-
       cluded on the example data sheets  shown in
       Figures 18-3 and 18-7. The audit analyses
       should agree with the audit concentrations
       within  10  percent.  When  available,  the
       tester may  obtain audit cylinders by con-
       tacting:  U.S.  Environmental  Protection
       Agency, Environmental Monitoring Systems

-------
                                                                                              Section  No.  3.16.10
                                                                                              Date  June  30,  19&8
                                                                                              Page  6
 Laboratory.  Quality  Assurance  Division
 (MD-77),  Research  Triangle  Park.  North
 Carolina  27711.  Audit cylinders obtained
 from a commercial gas manufacturer  may
 be used provided that (a) the gas manufac-
 turer  certifies  the  audit  cylinder  in  a
 manner similar  to the procedure described
 in 40 CFR Part 61. Appendix B. Method 106,
 Section 5.2.3.1, and (b) the gas manufactur-
 er obtains an independent analysis of the
 audit cylinders to verify this analysis. Inde-
 pendent analysis is defined  as an analysis
 performed by an individual other than the
 individual who performs the gas manufac-
 turer's  analysis,  while  using  calibration
 standards and analysis equipment different
 from those used for the gas manufacturer's
 analysis. Verification is complete and ac-
 ceptable   when  the  independent  analysis
 concentration is  within 5 percent of the gas
 manufacturer's concentration.
 7. Final Sampling and Analysis Procedure
   Considering safety  (flame hazards)  and
 the source conditions, select an appropriate
 sampling  and analysis procedure  (Section
 7.1, 7.2, 7.3, or 7.4). In situations where a hy-
 drogen flame Is a hazard and no Intrinsical-
 ly safe GC is suitable, use the flexible bag
 collection  technique or an adsorption tech-
 nique.  If  the source temperature Is below
 100'C,  and the  organic concentrations are
 suitable for the detector to be used, use the
 direct Interface method. If the source gases
 require dilution,  use a dilution Interface and
 either the bag sample or adsorption  tubes.
 The  choice between  these two  techniques
 will depend on the physical layout of the
 site,  the source temperature, and the stor-
 age stability of the compounds if collected
 in the bag.  Sample  polar compounds by
 direct Interfacing or dilution Interfacing to
 prevent sample loss by adsorption on  the
 bag.
   7.1 Integrated Bag Sampling and Analy-
 sis.
   7.1.1  Evacuated Container  Sampling Pro-
 cedure. In this procedure, the bags are filled
 by evacuating the rigid air-tight containers
 that hold the bags. Use a field  sample data
 sheet as shown in Figure 18-10. Collect trip-
 licate sample from each sample location.
  7.1.1.1  Apparatus.
  7.1.1.1.1  Probe.  Stainless  steel,  Pyrex
 glass, or Teflon tubing probe, according to
 the  duct  temperature, with 6.4-mm OD
 Teflon tubing of  sufficient length  to con-
 nect to the sample bag. Use stainless steel or
 Teflon unions to  connect probe and sample
 line.
  7.1.1.1.2  Quick  Connects. Male (2) and
 female (2) of stainless steel construction.
  7.1.1.1.3  Needle Valve.  To control gas
 flow.
  7.1.1.1.4  Pump. Leakless   Teflon-coated
 diaphragm-type pump or equivalent. To  de-
 liver at least 1  liter /tain.
  7.1.1.1.5   Charcoal Adsorption Tube. Tube
 filled with  activated  charcoal, with  glass
 wool plugs at  each end. to adsorb organic
 vapors.
  7.1.1.1.6   Flowmeter.  0  to  500-tnl  flow
range;  with   manufacturer's  calibration
curve.
   7.1.1.2  Sampling Procedure. To obtain a
 sample, assemble the sample train as shown
 in Figure 18-9. Leak check both the bag and
 the container.  Connect  the vacuum  line
 from the needle valve to the Teflon sample
 line from  the probe. Place the end of the
 probe at the  centroid of the stack,  or at a
 point no closer to the walls than 1  m, and
 start the pump with the needle valve adjust-
 ed to yield a flow of 0.5 liter/minute. After
 allowing sufficient time to purge the  line
 several times, connect the vacuum  line to
 the  bag, and  evacuate until the rotameter
 indicates no flow. Then position the sample
 and vacuum lines for sampling,  and begin
 the actual sampling, keeping the rate  pro-
 portional to the stack velocity. As a precau-
 tion, direct the gas exiting the rotameter
 away from sampling personnel. At the end
 of the sample period, shut off  the pump,
 disconnect the  sample line from the  bag,
 and disconnect  the vacuum line from the
 bag container. Record the source tempera-
 ture, barometric pressure, ambient tempera-
 ture, sampling  flow rate,  and initial  and
 final sampling time on the  data sheet shown
 in Figure 18-10. Protect the Tedlar bag and
 Its container from sunlight. When possible,
 perform the  analysis within  2 hours of
 sample collection.
  7.1.2  Direct Pump Sampling Procedure.
 Follow 7.1.1,  except place the pump  and
 needle valve between the probe and the bag.
 Use  a pump and needle valve constructed of
 stainless steel or some other material not af-
 fected  by  the stack gas.  Leak  check  the
 system, and  then purge  with  stack  gas
 before  the connecting to the  previously
 evacuated bag.
  7.1.3  Explosion Risk Area Bag Sampling
 Procedure. Follow 7.1.1 except replace  the
 pump  with  another  evacuated can  (see
 Figure  18-9a). Use this  method whenever
 there Is a possibility of an  explosion due to
 pumps, heated probes, or other flame pro-
 ducing equipment.
  7.1.4  Other Modified Bag Sampling Pro-
 cedures. In the  event that condensation is
 observed in the bag while collecting  the
sample and a direct interface system cannot
 be used, heat the bag during collection, and
maintain it at a suitably elevated tempera-
ture  during  all  subsequent  operations.
(Note:  Take care to leak check the system
prior to the dilutions so  as not to create a
potentially explosive atmosphere.) As an al-
ternative, collect the sample gas, and simul-
taneously dilute it In the Tedlar bag.
  In the first  procedure,  heat the box con-
taining the sample bag to  the source tem-
perature, provided the components  of  the
bag and the surrounding box can withstand
this temperature. Then transport the bag as
rapidly as  possible to the  analytical area
while maintaining the heating, or cover  the
box with an insulating blanket. In the ana-
lytical  area, keep the box heated to source
temperature until analysis.  Be sure that the
method of  heating the box and the control
for the heating circuit are  compatible with
the safety restrictions required in each area.

-------
                                                                                             Section  No.  3.16.10
                                                                                             Date  June  30,  1988
                                                                                             Page  7
  To use the second procedure, preflll the
Tedlar bag with a known quantity of Inert
gas. Meter the Inert gas into the bag accord-
Ing to the procedure for  the preparation of
gas  concentration  standards  of  volatile
liquid materials (Section 6.2.2.2). but elimi-
nate the midget Implnger section. Take the
partly  filled bag to the  source, and meter
the source  gas Into the bag through  heated
sampling lines  and a heated flowmeter,  or
Teflon  positive displacement pump. Verify
the dilution factors periodically through di-
lution  and analysis of gases of known con-
centration.
  7.1.5  Analysis of Bag Samples.
  7.1.5.1 Apparatus. Same as Section 5. A
minimum  of three gas  standards  are  re-
quired.
  7.1.5.2 Procedure.  Establish proper QC
operating conditions as described In Section
6.3, and record all data listed In Figure 18-7.
Prepare the GC  so that gas can be drawn
through the sample valve. Flush the sample
loop with gas from one of the three calibra-
tion  mixtures,   and   activate  the  valve.
Obtain  at least two chromatograms for the
mixture. The results are acceptable when
the peak areas from two consecutive injec-
tions agree to within 5  percent of their aver-
age. If they do not, run additional analyses
or correct  the analytical techniques until
this requirement is met. Then analyze the
other two calibration  mixtures In the same
manner. Prepare a calibration  curve as de-
scribed  In the same manner. Prepare a cali-
bration curve as described In Section 6.3.
  Analyze  the source  gas samples by con-
necting each bag  to the sampling  valve with
a piece  of Teflon tubing  identified for that
bag. Follow the specifications  on replicate
analyses specified for the calibration gases.
Record the data listed in  Figure 18-11. If
certain  items do not apply, use the notation
"N.A."  After all samples  have been ana-
lyzed, repeat the analyses of the calibration
gas mixtures, and generate a second calibra-
tion curve.  Use an average of the two curves
to determine the sample gas concentrations.
If the two calibration  curves differ by more
than 5 percent from their mean value, then
report the final  results  by comparison  to
both calibration curves.
  7.1.6  Determination of Bag Water Vapor
Content. Measure and record the ambient
temperature and barometric pressure near
the  bag. From  a water saturation vapor
pressure table, determine  and record the
water vapor  content  as  a  decimal  figure.
(Assume the relative humidity to be 100 per-
cent unless a lesser value Is known.) If the
bag has been maintained  at an elevated tem-
perature as described In Section 7.1.4. deter-
mine the stack gas water content by Method
4.
  7.1.7   Quality   Assurance.  Immediately
prior to the analysis of the stack gas sam-
ples, perform audit analyses as described In
Section 6.5. The audit analyses must agree
with the audit concentrations within  10 per-
cent. If the results are acceptable, proceed
with the analyses of the  source samples. If
they do not agree within 10 percent, then
determine  the reason  (or the  discrepancy,
and  take corrective action before proceed-
ing.
  7.1.8  Emission Calculations. From the av-
erage calibration curve described In Section

7.1.5., select the value of C. that corresponds
to the peak area. Calculate the concentra-
tion C, in ppm, dry basis, of each organic In
the sample as follows:
Cr  =
 WlFr
iTr  U-BWS)
                               •      -
C.=Concentration of the organic from the
   calibration curve, ppm.
P,=Reference  pressure,  the   barometric
   pressure or absolute sample loop pres-
   sure  recorded during calibration,  mm
   Hg.
T,=Sample loop temperature at the time of
   sample analysis, *K.
F,=Relative response factor (If  applicable.
   see Section 6.4).
P,=Barometric   or absolute sample loop
   pressure at time of sample analysis, mm
    Hg.
Tr=Reference temperature,  the  tempera-
   ture of the sample loop recorded during
   calibration/K.
8.,!= Water vapor content of the bag sample
   or stack gas, proportion by volume.
  7.2  Direct Interface Sampling and Analy-
sis Procedure. The direct Interface proce-
dure can be used provided that the moisture
content of the gas does not Interfere with
the analysis procedure, the physical require-
ments of the equipment can be  met at  the
site, and the source gas concentration is  low
enough that detector  saturation Is not a
problem. Adhere to all safety requirements
with this method.
  7.2.1   Apparatus.
  7.2.1.1  Probe.  Constructed of  stainless
steel,  Pyrex glass,  or Teflon tubing as  re-
quired by duct temperature,  6.4-mrn OD, en-
larged  at duct end to  contain  glass wool
plug. If necessary, heat the probe with heat-
Ing tape or a special heating unit capable of
maintaining duct temperature.
  7.2.1.2  Sample Lines. 6.4-mm  OD Teflon
lines, heat-traced to prevent condensation
of material.
  7.2.1.3  Quick   Connects.   To   connect
sample line to gas sampling valve on GC  In-
strument and to pump unit used to with-
draw  source gas.  Use  a quick  connect  or
equivalent on the cylinder or bag containing
calibration gas to allow connection of  the
calibration gas to the gas sampling valve.
  7.2.1.4  Thermocouple  Readout  Device.
Potentiometer  or digital thermometer,  to
measure source temperature and probe tem-
perature.
  7.2.1.5  Heated Gas Sampling  Valve.   Of
two-position,  six-port  design,   to  allow
sample loop to be purged with source gas or
to direct source gas into the  GC Instrument.
  7.2.1.6  Needle Valve. To control gas sam-
pling rate from the source.

-------
                                                                                              Section No.  3.16.10
                                                                                              Date  June  30,  198B
                                                                                              Page  8
   7.2.1.7 Pump. Leakless Teflon-coated dia-
  phragm-type pump or equivalent, capable of
  at least 1 liter/minute sampling rate.
   7.2.1.8 Flowmeter.  Of suitable range  to
  measure sampling rate.
   7.2.1.9 Charcoal Adsorber. To adsorb or-
  ganic vapor  collected from the source  to
  prevent exposure of personnel to source gas.
   7.2.1.10 Gas   Cylinders.   Carrier  gas
  (helium  or nitrogen), and oxygen and hy-
  drogen for a flame ionization detector (FID)
  if one is used. •
   7.2.1.11 Gas Chromatograph. Capable  of
  being moved into the field, with detector.
  heated gas sampling valve, column required
  to complete  separation of desired  compo-
  nents, and option for temperature program-
  ming.
   7.2.1.12 Recorder/Integrator.  To  record
  results.
   7.2.2  Procedure. To obtain a sample, as-
 semble the sampling  system as shown  in
 Figure 18-12. Make sure all connections are
 tight. Turn on the probe and sample line
 heaters.  As the temperature of the probe
 and heated line approaches the source tem-
 perature as indicated on the thermocouple
 readout device, control the heating to main-
 tain  a temperature of 0 to  3*C above  the
 source  temperature. While the probe and
 heated line are being heated,  disconnect the
 sample line  from the gas sampling valve,
 and attach the line from the  calibration gas
 mixture.  Flush the sample loop with cali-
 bration gas and analyze a portion of that
 gas. Record the results. After the calibra-
 tion gas sample has been flushed into the
 GC Instrument, turn the gas sampling valve
 to flush position, then reconnect the probe
 sample line to the valve. Place the Inlet of
 the probe at the centrold of the duct, or at a
 point no  closer to the walls than 1 m, and
 draw source gas into the probe, heated line.
 and sample loop.  After thorough flushing,
 analyze the sample using the same condi-
 tions as  for  the  calibration gas mixture.
 Repeat  the   analysis  on  an  additional
 sample. Measure the peak areas for the two
 samples, and if they do not agree to within  5
 percent of their mean value, analyze addi-
 tional samples until two consecutive analy-
 ses meet this criteria.  Record the  data.
 After  consistent   results  are   obtained,
 remove the probe from the source and ana-
 lyze  a  second calibration  gas mixture.
 Record this calibration data and the  other
 required  data  on the data sheet shown in
 Figure 18-11, deleting the dilution gas infor-
 mation.
  (NOTE: Take  care to draw all samples, cali-
 bration  mixtures,  and  audits through the
sample loop at the same pressure.)
  7.2.3 Determination  of Stack  Gas Mois-
ture Content. Use Method 4 to measure the
stack gas moisture content.
  7.2.4 Quality Assurance. Same as Section
 7.1.7.  Introduce  the  audit  gases  in the
sample  line  immediately  following the
 probe.
  7.2.5 Emission Calculations. Same as Sec-
 tion 7.1.8.
   7.3  Dilution   Interface  Sampling  and
 Analysis  Procedure.  Source  samples that
 contain a high concentration of organic ma-
 terials may require dilution prior to analysis
 to prevent saturating the GC detector. The
 apparatus required for this direct interface
 procedure is basically  the same as that de-
 scribed in the Section 7.2, except a dilution
 system is added between the heated sample
 line and the gas sampling valve. The appa-
 ratus  is arranged so that either  a  10:1 or
 100:1 dilution of the source gas can be di-
 rected to the  Chromatograph. A  pump of
 larger capacity is  also required,  and this
 pump must be heated and placed in the
 system between the sample line and the di-
 lution apparatus.
  7.3.1  Apparatus. The equipment required
 in addition to  that specified for the direct
 interface system is as follows:
  7.3.1.1  Sample  Pump.  Leakless Teflon-
 coated diaphragm-type that can withstand
 being heated to 120'C and deliver l.S liters/
 minute.
  7.3.1.2  Dilution Pumps. Two Model A-1SO
 Komhyr Teflon positive displacement type
 delivering 150 cc/minute. or equivalent.  As
 an option, calibrated flowmeters can be used
 in  conjunction  with  Teflon-coated  dia-
 phragm pumps.
  7.3.1.3  Valves.   Two  Teflon three-way
 valves, suitable for connecting, to 6.4-mm
 OD Teflon tubing.
  7.3.1.4  Flowmeters.  Two,  for  measure-
 ment of diluent gas, expected delivery flow
 rate to be 1,350 cc/mln.'
  7.3.1.5  Diluent Gas with Cylinders and
 Regulators. Gas  can  be nitrogen  or clean
 dry  air.  depending on  the nature of the
source gases.
  7.3.1.6 'Heated  Box.  Suitable for  being
 heated  to  120'C.  to   contain  the  three
 pumps, three-way  valves, and associated
 connections. The box  should be equipped
 with quick connect fittings to facilitate con-
 nection of: (1) The heated sample line from
 the probe, (2)  the  gas  sampling valve, (3)
 the calibration gas mixtures, and (4) diluent
 gas lines. A schematic diagram of the com-
 ponents and connections is shown in Figure
 18-13.
  (NOTE: Care must be taken to leak check
the system prior to the dilutions so  as not to
create a potentially explosive atmosphere.)
  The heated box shown in Figure 18-13 is
designed to receive a heated line from the
 probe. An optional design is to build a probe
 unit that attaches directly to the heated
box. In this way, the heated  box  contains
 the controls for the probe heaters,  or, if the
 box  is placed against the duct  being sam-
 pled, it may.be  possible  to eliminate the
 probe  heaters. In either case,  a heated
 Teflon line is used to connect  the heated
 box  to the gas sampling valve on the Chro-
 matograph.

-------
                                                                                             Section No.  3.16.10
                                                                                             Date  June  30,  1988
                                                                                             Page  9
   7.3.2  Procedure. Assemble the apparatus
 by connecting the  heated box. shown in
 Figure  18-13. between the heated sample
 line from  the probe and the gas sampling
 valve  on  the chromatograph.  Vent  the
 source gas from the gas sampling valve di-
 rectly to the charcoal filter, eliminating the
 pump and  rotameter.  Heat  the sample
 probe, sample line,  and  heated box. Insert
 the probe  and source thermocouple to the
 centrold of the duct, or to a point no closer
 to the walls  than  1  m. Measure the source
 temperature, and adjust all heating units to
 a temperature 0 to 3'C above this tempera-
 ture. If this  temperature Is above the safe
 operating temperature of the Teflon compo-
 nents, adjust the heating to maintain a tem-
 perature high enough to prevent condensa-
 tion  of  water  and  organic  compounds.
 Verify the  operation of the dilution system
 by analyzing a high concentration gas of
 known composition through either the 10:1
 or 100:1  dilution stages, as appropriate. (If
 necessary,  vary the flow of the diluent gas
 to obtain other dilution ratios.) Determine
 the concentration of the diluted calibration
 gas using the dilution factor and the  cali-
 bration curves prepared  in the laboratory.
 Record the pertinent data on the data sheet
 shown in Figure 18-11. If the data on the di-
 luted calibration gas are not within 10 per-
 cent  of  the  expected values, determine
 whether the chromatograph or the dilution
 system is In error, and correct It. Verify the
 OC operation using a low  concentration
 standard by diverting the  gas into  the
 sample loop,  bypassing the dilution system.
 If these analyses are not within acceptable
 limits, correct the dilution system to provide
 the desired dilution factors. Make this cor-.
 rectlon  by  diluting  a high-concentration
 standard gas  mixture to adjust  the dilution
 ratio as required.
  Once the dilution system and  GC oper-
 ations are  satisfactory,  proceed  with the
 analysis of  source gas. maintaining the same
 dilution settings as used for the standards.
 Repeat the analyses  until two consecutive
 values do not vary by more than S percent
 from their mean value are obtained.
  Repeat the analysis of the calibration gas
 mixtures to  verify  equipment operation.
 Analyze the two field audit samples using
 either the dilution system, or directly  con-
 nect to the gas sampling valve  as required.
 Record all data and report the results to the
 audit supervisor.
  7.3.3 Determination of  Stack Oas Mois-
 ture Content. Same as Section 7.2.3.
  7.3.4 Quality Assurance. Same as Section
 7.2.4.
  7.3.5 Emission Calculations. Same as Sec-
 tion 7.2.5, with the  dilution factor applied.
  7.4  Adsorption Tube Procedure (Alterna-
 tive Procedure). It  is suggested that the
 tester refer to the National Institute of Oc-
 cupational  Safety  and  Health  (NIOSH)
 method for the particular organics to be
 sampled.  The principal Interferent will be
 water vapor.  If water vapor is present at
 concentrations above 3 percent,  silica gel
 should be  used In  front  of  the charcoal.
 Where more than one compound is present
 In the emissions, then develop relative ad-
sorptlve capacity information.
  7.4.1  Additional  Apparatus. In addition
 to  the  equipment  listed in  the NIOSH
 method for the particular organic(s) to be
 sampled, the following items (or equivalent)
 are suggested.
  7.4.1.1 Probe  (Optional).   Boroslllcate
 glass or stainless steel, approximately 6-mm
 ID, with a heating system If water conden-
 sation Is a problem, and a filter (either in-
 stack or out-stack heated to stack tempera-
 ture) to remove paniculate matter. In most
 Instances, a plug of glass wool Is a satisfac-
 tory filter.
  7.4.1.2 Flexible Tubing. To connect probe
 to adsorption tubes. Use a material that ex-
 hibits minimal sample adsorption.
  7.4.1.3 Leakless Sample Pump. Flow con-
 trolled, constant rate pump, with a set of
 limiting (sonic) orifices to provide  pumping
 rates from approximately 10 to 100 cc/mln.
  7.4.1.4 Bubble-Tube Flowmeter.  Volume
 accuracy within ± 1 percent, to  calibrate
 pump.
  7.4.1.S Stopwatch. To time sampling and
 pump rate calibration.
  7.4.1.6 Adsorption Tubes. Similar to ones
 specified by NIOSH. except the amounts of
 adsorbent per primary /backup sections are
 800/200 mg for charcoal tubes and  1040/260
 mg for silica gel tubes. As an alternative.
 the tubes may contain a porous polymer ad-
 sorbent such as Tenax OC or XAD-2.
  7.4.1.7 Barometer. Accurate to 5 mm Hg.
 to  measure  atmospheric pressure during
 sampling and pump calibration.
  7.4.1.8  Rotameter. 0  to 100 cc/mln,  to
 detect changes in flow rate during sampling.
  7.4.2  Sampling and  Analysis. It Is sug-
 gested that the tester  follow the sampling
 and analysis  portion  of the respective
 NIOSH  method  section  entitled   "Proce-
 dure." Calibrate the pump and limiting ori-
 fice flow rate through adsorption tubes with
 the bubble  tube flowmeter before sampling.
 The sample system can be operated as a "re-
circulating  loop" for this operation. Record
 the ambient temperature  and barometric
 pressure. Then, during sampling, use the ro-
 tameter to verify that the pump and orifice
sampling rate remains constant.
  Use a sample probe. If required, to obtain
 the sample at the centroid of the duct, or at
 a point no closer to the walls than 1 m. Min-
 imize the length of flexible tubing between
 the probe and adsorption tubes. Several ad-
 sorption tubes can be connected in series. If
 the extra adsorptlve capacity is needed. Pro-
 vide the gas sample to the sample system at
 a pressure sufficient for the limiting orifice
 to  function as a sonic  orifice. Record the
 total  time  and sample flow  rate  (or the
 number of pump strokes), the barometric
 pressure, and ambient temperature. Obtain
 a total  sample volume commensurate with
 the expected concentration(s) of the volatile
 organic(s)   present,   and   recommended
 sample  loading factors (weight sample per
 weight adsorption media). Laboratory tests
 prior  to actual sampling may  be necessary
 to  predetermine this volume. When  more
 than one organic is present in the emissions,
 then develop relative adsorptlve capacity in-
 formation.  If water vapor is present in the
 sample  at concentrations above 2 to 3 per-
 cent, the adsorptlve capacity may be severe-

-------
                                                                                              Section No.  3.16.10
                                                                                              Date  June  30,  1988
                                                                                              Page  10
 ly reduced. Operate the gas chromatograph
 according to the manufacture's instructions.
 After  establishing  optimum  conditions.
 verify  and  document  these   conditions
 during all  operations.  Analyze  the audit
 samples (see Section 7.4.4.3), then the emis-
 sion samples. Repeat the analysis of each
 sample until  the relative deviation of two
 consecutive injections does not exceed 5 per-
 cent.
   7.4.3  Standards  and  Calibration.   The
 standards can be prepared according to the
 respective NIOSH method. Use a minimum
 of three different standards; select the con-
 centrations to bracket the expected average
 sample concentration. Perform the calibra-
 tion before and  after  each  day's sample
 analyses. Prepare the calibration curve by
 using the least squares method.
   7.4.4  Quality Assurance.
   7.4.4.1  Determination of Desorption Effi-
 ciency. During  the  testing program, deter-
 mine the desorptlon efficiency In the ex-
 pected sample concentration range  for each
 batch of  adsorption media to be used.  Use
 an Internal standard. A minimum desorp-
 tlon  efficiency  of 50  percent shall be ob-
 tained. Repeat  the desorptlon  determina-
 tion until the relative deviation of two con-
 secutive determinations does  not  exceed 5
 percent. Use  the average desorptlon effi-
 ciency of these  two consecutive determina-
 tions for the correction specified in Section
 7.4.4.5.  If the desorptlon efficiency of the
 compound(s)  of  Interest is  questionable
 under actual sampling conditions, use of the
 Method of Standard Additions may be help-
 ful to determine this value.
   7.4.4.2  Determination of Sample Collec-
 tion Efficiency. For the source samples, ana-
 lyze the primary and backup portions of the
 adsorption tubes separately.  If the backup
 ponton exceeds 10 percent  of  the total
 amount (primary  and backup), repeat  the
 sampling with a  larger sampling portion.
  7.4.4.3  Analysis   Audit.   Immediately
 before the sample analyses, analyze the two
 audits in accordance with Section 7.4.2. The
 analysis audit shall agree with the  audit
 concentration within 10 percent.
  7.4.4.4  Pump  Leak Checks  and  Volume
 Flow Rate Checks. Perform both  of these
 checks immediately after sampling  with all
 sampling  train  components in place.  Per-
 form all leak checks according to the manu-
 facturer's  instructions, and record  the re-
 sults. Use the  bubble-tube  flowmeter  to
 measure the pump volume flow rate with
 the orifice used in the test sampling, and
 the result. If it has changed by more than 5
 but less than 20 percent, calculate an aver-
 age flow rate for the test. If the  flow rate
 has changed by more than 20 percent, reca-
 librate the pump and repeat the sampling.
  7.4.4.5  Calculations.  All  calculations can
be performed  according to the respective
 NIOSH method.  Correct all sample volumes
 to standard conditions. If a sample dilution
system has been used,  multiply the results
by the appropriate dilution ratio. Correct all
results by dividing by  the desorptlon effi-
ciency (decimal  value).  Report  results as
ppm by volume, dry basis.
 ' 7.5  Reporting of Results. At the comple-
 tion of the  field analysis  portion of the
 study, ensure that the data sheets shown in
 Figure 18-11  have been completed. Summa-
 rize this data on the data sheets shown  In
 Figure 18-15.
 8. Bibliography
   1. American Society for Testing and Mate-
 rials. Ci  Through C. Hydrocarbons in the
 Atmosphere   by  Gas  Chromatography.
 ASTM D 2820-72, Part 23. Philadelphia. Pa.
 23:950-958. 1973.
   2. Corazon, V. V. Methodology for Collect-
 ing and Analyzing Organic  Air Pollutants.
 U.S.   Environmental  Protection  Agency.
 Publication No.  EPA-600/2-79-042. Febru-
 ary 1979.
   3. Dravnieks, A.,  B. K. Krotoszynski,  J.
 Whitfield, A. O'Donnell, and T. Burgwald.
 Environmental  Science  and  Technology.
 5(12):1200-1222. 1971.
   4. Eggertsen. F. T., and F. M. Nelsen. Gas
 Chromatographlc Analysis of Engine Ex-
 haust and Atmosphere.  Analytical  Chemis-
 try. 30(6): 1040-1043. 1958.
   5. FeairheUer,  W. R.,  P. J. Marn. D.  H.
 Harris, and D. L. Harris. Technical Manual
 for Process Sampling Strategies for Organic
 Materials. U.S. Environmental Protection
 Agency. Research Triangle Park. NC. Publi-
 cation No.  EPA 600/2-76-122. April  1976.
 172 p.
   6. FR. 39 FR 9319-9323. 1974.
   7. FR, 39 FR 32857-32860. 1974.
   8.  FR,  41  FR  23069-23072 and  23076-
 23090. 1976.
   9. FR, 41 FR 46569-46571. 1976.
   10. FR, 42 FR 41771-41776. 1977.
   11. Fishbein. L. Chromatography of Envi-
 ronmental Hazards, Volume II. Elsevier Sci-
 entific Publishing Company.  NY. NY. 1973.
   12. Hamersma. J. W.,' S. L.  Reynolds, and
 R. F.  Maddalpne.  EPA/IERL-RTP  Proce-
 dures Manual: Level 1 Environmental As-
 sessment.  U.S.  Environmental  Protection
 Agency. Research Triangle Park. NC. Publi-
 cation No. EPA  600/276-160a. June  1976.
 130 p.
  13. Harris, J. C.. M. J. Hayes. P. L. Levins.
 and D. B. Lindsay. EPA/IERL-RTP Proce-
 dures for Level 2 Sampling and Analysis of
 Organic Materials. U.S. Environmental  Pro-
 tection  Agency.  Research Triangle Park,
 NC. Publication No. EPA 600/7-79-033.  Feb-
 ruary 1979. 154 p.
  14.  Harris. W.  E.. H.  W. Habgood.  Pro-
 grammed  Temperature Gas  Chromatogra-
 phy. John Wiley & Sons, Inc. New York.
 1966.
  15.  Intersociety  Committee. Methods of
Air  Sampling  and   Analysis.   American
Health Association. Washington, DC. 1972.
  16. Jones, P. W., R. D. Grammar, P. E.
Strup, and T. B. Stanford. Environmental
Science and Technology. 70:806-810. 1976.
  17. McNair Han Bunelli, E.  J. Basic  Gas
Chromatography.  Consolidated  Printers.
Berkeley. 1969.
  18. Nelson, G. O. Controlled Test  Atmos-
pheres,  Principles  and  Techniques.   Ann
Arbor Ann Arbor Science Publishers. 1971.
247 p.

-------
                                                                                          Section  No.  3.16.10
                                                                                          Date  June  30,  1988
                                                                                          Page  11
  19. NIOSH Manual of Analytical Methods.
Volumes 1, 2, 3. 4. 5, 6, 7. U.S. Department
of Health and Human Services National In-
stitute for Occupational Safety and Health.
Center for Disease Control. 4676 Columbia
Parkway,  Cincinnati.  Ohio  45226.  April
1977-August  1981. May be available from
the Superintendent of Documents, Govern-
ment  Printing  Office, Washington,  DC
20402. Stock Number/Price: Volume 1—017-
033-00267-3/$13, Volume 2-017-033-00260-
6/$ll,    Volume   3—017-033-00261-4/S14,
Volume  4-017-033-00317-3/S7.25, Volume
5-017-033-00349-1/S10, Volume 6-017-033-
00369-6/$9, and  Volume 7-017-033-00396-
5/$7.  Prices  subject  to  change.  Foreign
orders add 25 percent.
  20. Schuetzle, D.. T. J. Prater, and S. R.
Ruddell.  Sampling and Analysis of Emis-
sions from Stationary Sources; I. Odor and
Total Hydrocarbons. Journal of the Air Pol-
lution  Control Association. 25(9):925-932.
1975.
  21. Snyder. A.  D.. F. N.  Hodgson, M. A.
Kemmer and J.  R. McKendree. Utility of
Solid Sorbents for Sampling Organic Emis-
sions from Stationary' Sources. U.S. Envi-
ronmental Protection Agency. Research Tri-
angle Park. NC Publication No. EPA 600/2-
76-201. July 1976. 71 p.
  22.  Tentative  Method   for  Continuous
Analysis of Total Hydrocarbons in the At-
mosphere. Intersociety Committee. Ameri-
can Public Health Association. Washington.
DC 1972. p. 184-186.
  23. Zwerg, G.. CRC Handbook of Chroma-
tography, Volumes I and II. Sherma. Joseph
(ed.). CRC Press. Cleveland. 1972.

-------
I.  Nut of aonpany
   Mdrtss

   Contacts
 Oat*
 Phone
   Process to be sampled
   Duet or vint to be sampled_
II. Process description
   lUw material
   Products
   Operating cycle
       Check:  Batch	Continuous
       Timing of batch or cycle	
       Best time to test 	
.Cyclic
           Figure 18-1.  Preliminary survey data sheet.
                                                               Component! to be analyzed
                                                    expected concentration
                                                               Suggested chromatographic column
Column flow rat*      ml/min   Bead  pressure      gon Hg
Column temperatore<
     Isothermal            *C
     Programmed from      *C to      *C at 	•C/min
Injection port/sample) loop temperature       *C
Detector  temperature        *C
Detector  flow ratea: Hydrogen       ml/min.
                                head  pressure      mm Hg
                    Air/Oxygen       ml/min,
                                head  preaaure      mn Hg
Chart speed __^______> inches/minute
Compound  data*
     Compound     Retention time     Attenuation
                                                                 Figure 18-2.  Chromatographic  conditions  data sheet.
                                                                                T) O W
                                                                                03 SO 0)
                                                                                Ot} rt D
                                                                                (D (D rt
                                                                                     p.
                                                                                I-1 t-i O
                                                                                M C 3
                                                                                   £3
                                                                                   n> -z.
                                                                                                                            uo
                                                                                                                            o
                                                                                                                               OJ
                                                                                                                            U) CT\
                                                                                                                            CD*'
                                                                                                                            CO H-
                                                                                                                               O

-------
                  Preparation of Standards In Tedlar  Bags
                           and Calibration Curve
                                                        Standards
                                               Mixture   Mixture
Standards Preparation Data:
  Organic: __^^_^_
  Bag number or Identification
  Dry gas meter calibration factor
  Final meter reading (liters)
  Initial meter reading (liters)
  Mete red volume (liters)
  Average meter temperature CK)
  Average meter pressure,  gauge (mm Hg)
  Average atmospheric pressure (mm Hg)
  Average meter pressure,  absolute (mm Hg)
  Syringe temperature CK)
    (Section 6.2.2.1)
  Syringe pressure,  absolute (mm Hg)
    (Section 6.2.2.1)
  Volume of gas In syringe (ml)
    (Section 6.2.2.1)
  Density of liquid  organic (g/ml)
    (Section 6.2.2.2)
  Volume of liquid In syringe ({!)
    (Section 6.2.2.2)

GC Operating Conditions:

  Sample loop volume (ml)
  Sample loop temperature  CO
  Carrier gas flow rate (ml/mln)
  Column temperature
    Initial  CO
    Rate change CC/m1n)
    Final CO

Organic Peak Identification and
  Calculated Concentrations:

  Injection time  (24-hr clock)-
  Distance to peak (cm)
  Chart speed (cm/mln)
  Organic retention  time (mini
  Attenuation factor
  Peak height (mm)
  Peak area  (mm2)
  Peak area  x attenuation  factor (rm2)
  Calculated concentration (ppm)
    (Equation 18-3 or 18-4)

  Plot peak  area  x attenuation factor against calculated concentration
  to obtain calibration curve.

        Figure  18-3.   Standards prepared In Tedlar  bags
                      and  calibration curve.
                           Flow»et*r Calibration
Flowneter number or Identification
Flowmeter type
Calibration device Ix):Bubble meter_
Readings at laboratory conditions:
  Laboratory temperature (T)a[))
  Laboratory barometric pressure
Flow data:
             Flowmeter
                                       Sp1rometer_
                                                         Wet  test meter
                                                     Hg
                                               Calibration device
  reading   I
(as narked)!
               temp.
               CK)
1  pressure
((absolute)
Time I             I
(mln) I gas  volume*  I flow  rateb
   a * Volume of gas measured by calibration device, corrected  to  standard
        conditions (liters).

   b • Calibration device gas volume/time.

Plot flowmeter reading against flow rate (standard conditions), and draw  a
smooth curve.  If the flowmeter being calibrated  1s a rotaaeter or other
flow device that Is viscosity dependent, It  may be necessary to generate  a
'fully* of calibration curves that cover the operating pressure and
temperature ranges of the flowneter.

While the following technique should  be  verified  before application. It may
be possible to calculate flow rate readings  for rotameters at standard
conditions Q$td « follows:
                       Qstd.  • Qlab
                                 /760M

                                 VTalT
                                         * "
                                                 1/2
         Flow rate
  (laboratory conditions)
                                                     Flow rate
                                                (standard conditions)
            Figure  18-4.   Flowneter calibration.
                                                                            T3 U C/>
                                                                            0 pj  CD
                                                                            (jq rt O
                                                                            CD CD  rt
                                                                                  H-
                                                                            H" C_, O
                                                                            OJ C  3

                                                                               CD  "Z
                                                                                  O
                                                                                  00
                                                                                  OO (

-------
                                                                      Section No. 3.16.10
                                                                      Date  June  30,  1988
                                                                      Page  14
     COMPONENT
        CAS
      CYLINDER
                                      \
                             CALIBRATED ROTAMETERS
                               WITH ROW CONTROL
                                    VALVES
       DILUENT
         CAS
       CYLINDER
                  Figure 18-5.  Single-stage calibration gas dilution systea.


                                    HIGH
                                CONCENTRATION
                                   WASTE
                                     I.
—NEEDLE VALVES
          ROTAMETERS ~
                        m

                m
     LOW
•CONCENTRATION
     GAS
                                          PRESSURE
                                'PR '      REGULATOR
                            DILUENT AIR
             DILUENT AIR
    PURE SUBSTANCE OR
PURE SUBSTANCE/N2 MIXTURE
                    Figure 18-6.  Two-stage dilution apparatus.

-------
                                                                           Section  No.   3.16.10
                                                                           Date June  30,   1988
                                                                           Page 15
          Preparation of Standards by Dilution of Cylinder Standard

 Cylinder standard:  Organic 	  Certified concentration	  ppn

 Standards Preparation Data:                            Date 	
                                      Mixture 1     Mixture 2     Mixture 3
   Standard gas  flowmeter reading
   Diluent gas  flowraeter reading
   Laboratory temperature CK)
   Barometric pressure (mm Hg)
   Flowmeter gage pressure (mm Hg)
   Flow rate cylinder gas at
    standard conditions (ml/mln)
   Flow rate diluent gas at
    standard conditions (ml/mini
   Calculated concentration (ppn)

             Stage 2 (If used)

   Standard  gas flowmeter reading
   Diluent gas flowmeter reading
   Flow rate stage 1 gas at
    standard conditions (ml/mln)
   Flow rate diluent gas at
    standard conditions (ml/mini
   Calculated concentration (ppn)

 GC Operating Conditions:
   Sample loop volume (ol)
   Sample loop temperature  (*C)
   Carrier gas flow rate (ml/mln)
   Column temperature:
        Initial  CO
       Program rate  CC/min)
       Final CO

Organic Peak Identification and
Calculated Concentrations:
   Injection time (24-hr clock)
  Distance to  peak (cm)
  Chart speed  (cm/mln)
  Retention time (m1n)
  Attenuation  factor
  Peak area (mm2)
  Peak area x  attenuation factor
Plot peak  area  x attenuation factor against calculated concentration  to
obtain calibration curve.

        Figure  18-7.  Standards prepared by dilution of cylinder standard.

-------
                                                            Section  No. 3.16.10
                                                            Date June 30,  19§8
                                                            Page 16
                          BOILING -*
                           WATER
                            BATH
                                            SYRINGE
                                             SEPTUM
                                           14- MIDGET
                                              IMPINGER
                                    HOTPLATE
                    NITROGEN
                    CYLINDER
                Figure 18-8.  Apparatus for preparation of liquid Materials.
                                                                        VENT
        STACK
        WAIL
    \
 REVERSE
 0"IIYPE
P1TOTTUBE
            PITOT MANOMETER
                        RIGID LEAKPROOF CONTAINER
                Figure 18-9.   Integrated bag sampling train.

-------
                                                           Section No.  3.16.10
                                                           Date June  30,  1988
                                                           Page 17
                           WC Tubing
                          FloMHtter
                                                                MractlMl
                                                                HtHIt falw
Mr Tight Stetl On* ,








; "*"•••.
' t
• *
• ^. *
'•% ;
\ /
j
* •
« /








-VfVf 	
CVBCVBtAQ StGCI
Droi













             Figure 18-9a. Explosion risk g«s sailing «ethod.
Plant.
S1t«
   Date
                         Sample 1
Seurci twperatura (*C)
Barantrlc prissurt(nn Hg)_

Ambltnt tcop«riturt (*C) _

Sanplt flow r»te(appr.)  _

Bag nunber

Start tine

Finish tine
Sample 2
Sample  3
         Figure 18-10.   Field sample data sheet - Tedlar
                     bag collection method.

-------
                                                                 Section  No.  3.16V10
                                                                 Date  June 30,
                                                                 Page  18
 riant	;	Date_
 Location
 1.   General information
           Source temperature (*C)
           Probe temperature (•«
           Ambient temperature (*C)
           Atmospheric pressure (am)
           Source pressure ('Eg)
           Absolute source pressure (an)
           Sampling rate (liter/min)
           •ample loop volume (ml)
           Sample loop temperature  (*C)
           Columnar temperature i
                initial (*C)/tiM (min)
                Program nu  CC/min)
                Final (*C)/tlM
           Carrier gas flow rate (ml/min)
           Detector temperature (*C)
           Injection time (24-hour buls)
           Chart speed (mm/min)

           Dilution gas flow rate (ml/min)
           Dilution Cas used (symbol)
           Dilution ratio
                Figure 18-11. Field tnilysls dtti shttts.

2.  Tield Analysis Data - Calibration Gas
    tan MO.              Time
         ments   Area   Attenuation   A « A factor   Cone,  (mm)
    JU-n He.	    Ti»e	
    Cempcnents    Area   Attenuation   A K A Factor   Cone, (ppm)
   »un Mo.	   Time	
   Components   Area   Attenuation   A « A Factor   Cone,  (ppm)
             Figure 18-11 (tontlmttd).  Field antlytls dtU »httt».

-------
                                                                  Section No.  3.16.10
                                                                  Date  June 30,  1988
                                                                  Page  19
TC
KAOOUT
	 M


T C RMKU1
OR
ONitoun


         UUIUlllUUUI.
. I1BULATKM
                                                  mm ots
                                                 SMVUNCVM.W
                                                    met
                     Figure 18-52.  Direct Interface sampling syste*.
                         Vent to Charcoal  Adsorbers
Heated Line
From Probe
t




t
=X»E
Quick
Connect







-Q~
Source
Gas Pump
i




IbU
cc/M1r
Pun
—t-
1.5 L/M1n "

i
10:1 100:1
'VJ^




'-&
)



-4
3-Uay Valves
In 100:1
Position '
^—^
-i
)


Quick Connects
To Gas Sample
Valve


J 150 cc/MIn
i Pump
Check Valve I
' Quick Connects 1
> For Calibration t


||





                                                                       FloMneters
                                                                       (On Outside
                                                                        Of Box)

                                                                       Flow Rate Of
                                                                       1350 cc/MIn
                      Heated Box at 120 C Or Source Temperature
 Figure 18-13.   Schematic  diagram of the heated box required
               for dilution of sample gas.

-------
                                                                                  Section  No.  3.1640
                                                                                  Date  June   30,  1988
                                                                                  Page  20
   OAUOUI OBOAKIC Bummo AJTD AHALTBU
                 CHECKLIST

      (Respond with Initials or number u                        £2JK
                appropriate)                                     i
                                              1. General
                                               Information:
                                                Source-
                                                  temperature
                                     Da««         CO
 1. Praumrdau:                                 Probe
       A. Grab ample collected...   D	        temperature
       B. Onto ample analyzed                    i—hi—i.
         for compoaltlan	a               Ambient
                  CO-
                 mblex
                  temperature


a        D
             Method OC	D        D        ....
                                a        D        'c>
                                                Atmospheric

       C.  OC-FTD analysis per-                      {mTa?)
        (onned	a ._
                                                Source
 2. laboratory calibration data:
       A. Calibration curvea pre-	

             Number of oompo-                     (ml/mln)-
              nenU	a        O      Sample loop
                                                 volume (ml).. _
                                                Sample loop
                                                 temperature
                                                 CO-
                                                Sample
                                                 collection
                                                 time (24-hr
                                                 basis)...
                                                Column
                                                 tempera.
            OK  obtained  for                      tun:
              field work..——...   O ,,,,	-          i^mai
 3. Sampling, procedures:            ,                     CO.—
      A. Method:                                    Prosram
            Bas ample	O       D           rate CC/
            Direct Interface	   O       D           mini..
            Dilution Interface—   D       O          Final CC)...
      B. Number of aampla col-                     Carrier iss
        leeud ___...	_..   1O ..._..__	        low rate
 4. Field analysis:                                   (ml/mln)	
      A.  Total  hydrocarbon                     Detector
        analysts performed——   O 	        temperature
      B. Calibration curve pre-
                                                 CO..
            *"'""" »"""' *»•«-                    |-tl»rt ••..•1
        pared _________   Q —____..      Chart 
-------
                                                                                          Section No.  3.16.10
                                                                                          Date  June 30,  1988
                                                                                          Page  21
   APPENDIX C—QUALITY ASSURANCE
              PROCEDURES *

  Procedure 1—Determination of Adequate
     Chromatographic Peak Resolution
  In this method of dealing with resolution,
 the  extent to which  one Chromatographic
 peak overlaps another is determined.
  For convenience, consider the range of the
 elution curve of each  compound as running
 from -2cr to +20-. This range is used in
 other  resolution criteria,  and it  contains
 95.45 percent of the area of a normal curve.
 If two peaks are separated by a known dis-
 tance,  b, one can determine the fraction of
 the  area of one curve that lies within the
 range of the other. The extent to which the
 elution curve of a contaminant compound
 overlaps the curve of a  compound that is
 under  analysis is found by integrating the
 contaminant curve over the limits b-2cr. to
 b-f-2cr,, where a-, is the standard deviation of
 the sample curve.
  This calculation can be simplified in sever-
 al ways.  Overlap  can be determined for
 curves of unit area; then actual areas can be
 introduced.  Desired integration can  be re-
 solved  into two integrals  of the normal dis-
 tribution function for which there are con-
 venient calculation programs and tables. An
 example would be Program 15 In Texas In-
 struments  Program  Manual  ST1,  1975,
 Texas   Instruments,  Inc.,  Dallas,  Texas
 75222.
  In judging the suitability of alternate GC
 columns or the effects of altering Chromato-
 graphic conditions, one can employ the area
 overlap as the resolution parameter with a
 specific maximum permissible value.
  The use  of Gaussian functions to describe
 Chromatographic elution  curves  is  wide-
 spread.  However, some elution curves are
 highly   asymmetric.  In  cases  where  the
 sample  peak is followed by a contaminant
 that has a leading edge that rises  sharply
 but the curve then tails off. it may be possi-
 ble  to  define an effective  width for tc as
 "twice the distance from the leading edge to
a perpendicular line through the maxim of
the contaminant curve, measured along a
perpendicular bisection of that line."
 Procedure 2—Procedure for Field Auditing
               GC Analysis

  Responsibilities of audit supervisor  and
analyst at the source sampling site include
the following:
  A. The audit supervisor verifies that audit
cylinders are stored in  a safe location  both
before and after the audit to  prevent van-
dalism.
  B. At the beginning and conclusion of the
audit,  the analyst records  each cylinder
number and pressure. An audit cylinder is
never  analyzed when  the  pressure drops
below 200 psi.'
  C. During the audit, the analyst performs
a minimum of two consecutive analyses of
each audit cylinder gas. The audit must be
conducted to coincide with the analysis of
source test samples,  normally  immediately
after  GC calibration and prior to sample
analyses.
  D. At the end of audit analyses,  the audit
supervisor requests the calculated concen-
trations from the analyst and compares the
results with the actual audit concentrations.
If each measured concentration agrees  with
the respective actual concentration  within
±10 percent, he directs  the analyst to begin
analyzing source samples. Audit supervisor
judgment and/or supervisory policy deter-
mine action when agreement is not within
±10 percent. When a consistent bias  in
excess  of 10 percent is found, it may be pos-
sible to proceed with the sample analysis.
with a corrective factor to be applied to the
results at a later time. However,  every at-
tempt should be made to locate the cause of
the discrepancy, as it  may be misleading.
The audit supervisor records each cylinder
number, cylinder pressure (at the end of the
audit), and  all calculated  concentrations.
The individual being  audited must   not
under any circumstance be told actual audit
concentrations until  calculated concentra-
tions have been submitted to the  audit su-
pervisor.
    CFR-61, Appendix C,  Quality  Assurance  Procedures,  July  1,  198?,  pages 157  -  161.

-------
                                                                                                  Section  No.  3.16.10
                                                                                                  Date  June  30,
                                                                                                  Page  22
        b»2o
       b-2o,                b-2as            b+2os
                           ~°T    '        ~
 The following calculation steps are required:*
 1.  2os = tt/A In 2
 2.  oc = tc/2V2 In 2
 3.  x, = (b-2of)/oc
 4.  x, = (b+2oj)/oc
                       dx
6.   Q(x2)
           m
                       'dx
7-   L
                *»
              - Q(XI)
    Ao •
9.   Percentage overlap = A  x  100 ,
where:
     A  = Area of the sanple peak of Interest determined by electronic Inte-
         gration or by the formula AS = hftt-
     A  * Area of the contaminant peak, determined  in the same manner as A$.
     b « Distance on the chrooatographlc chart that separates the maxima of
         the two peaks.
     H  = Peak height of the sanple  compound of Interest, measured from the
         average value of the baseline to the maximum of the curve.
     ts =• Width of sample peak of Interest at 1/2 peak height.
     tc = Width of the contaminant peak at 1/2 of peak height.
    o  - Standard deviation of the  sanple compound of Interest elutlon
         curve.
    a  « Standard deviation of the  contaminant elutlon curve.
 Q(X|) = Integral  of the normal  distribution function from x, to Infinity.
 Q(XJ) = Integral  of the normal  distribution function from x2 to Infinity.
     I  = Overlap Integral.
    AO * Area  overlap fraction.
   *In s»st Instances,  Q(x2) Is very snail and nay be neglected.
                                                                                              FIELD AUDIT REPORT
                                                                                   Part A.—To be filled out by  organization
                                                                                 supplying audit cylinders.
                                                                                   1.  Organization supplying audit sample(s)
                                                                                 and  shipping address
                                                                                   2.  Audit  supervisor,  organization,  and
                                                                                 phone number
                                                                                   3.  Shipping instructions:  Name, Address.
                                                                                 Attention
                                                                                   4.  Guaranteed  arrival  date  for  cylin-
                                                                                   5.  Planned  shipping  date   for  cylin-
                                                                                   6.  Details on audit cylinders  from  last
                                                                                 analysis

a. Date of last analysis 	
b Cytinder number
c. Cylinder pressure, psi 	
d. Audit gas(es)/balance gas 	 -.

f. Cylinder construction 	

Low cone.







High cone.







                                                                                  Part B.—To be filled out by audit supervi-
                                                                                sor.
                                                                                  1. Process sampled-
                                                                                  2. Audit location-
                                                                                  3. Name of individual audit-
                                                                                  4. Audit date	
                                                                                  5. Audit results:




d. Measured concentration, ppm Injection
#1" Injection #2' Average .. 	
e. Actual audit concentration, ppm (Part A.
So) 	
1. Audit accuracy:1
High Cone Cylinder .... 	
Percent' accuracy -
Measured Cone. - Actual Cone.
Actual Cone.

Low
cone.
cylin-
der










High
cone.
cylin-
der










                                                                                  1 Results  of two consecutive  infections  that  meet  the
                                                                                sample analysis cntena of the test method.
                                                                                [47 PR 39178. Sept. 7. 1982]

-------
                                                          Section No. 3-l6.ll
                                                          Date June 30, 1988
                                                          Page 1
 11.0   REFERENCES
 1.    Method  18 -  Measurement  of Gaseous Organic Compound  Emissions  by  Gas
      Chromatography.   Federal  Register. Volume 48, No. 202, October 1-8, 1983,
      page 48344.

 2.    Amendments to Method 18.   Federal Register. Volume 49, No. 105, May  30,
      1984,  page 22608.

 3.    Miscellaneous Clarifications   and Addition of Concentration Equations to
      Method 18. Federal Register.  Volume 52, No. 33, February 19, 198?, page
      5105-

 4.    Stability  of Parts-Per-Million  Organic  Cylinder Gases and Results of
      Source  Test   Analysis  Audits,  Status  Report #8.  U. S. Environmental
      Protection Agency Publication No. EPA-600/2-86-117, January 198?.  Also
      available from NTIS as  Publication No. PB 8?-l4l46l.

 5.    Traceability   Protocol  for Establishing True Concentration of Gases Used
      for  Calibration and Audits   of  Continuous  Source  Emission  Monitors
      (Protocol  No.   1).  Section  3-0.4, Quality Assurance Handbook, Volume
      III, Stationary  Source  Specific Methods, U. S. Environmental  Protection
      Agency Publication No.  EPA-600/4-77-027b, June 15, 1978.

 6.    Methanol,  Method  2000.   NIOSH Manual of Analytical Methods, Volume  2,
      Third Edition, U.  S. Department  of Health and Human Services, February
      1984.

 7.    Alcohols  I,  Method 1400.  NIOSH Manual of Analytical Methods, Volume 1,
      Third Edition, U.  S. Department  of Health and Human Services, February
      1984.

 8.    Alcohols II,  Method 1401.  NIOSH Manual of Analytical Methods, Volume 1,
      Third Edition, U.  S. Department  of Health and Human Services, February
      1984.

 9.    Hydrocarbons, BP 36 - 126° C,  Method 1500.  NIOSH Manual  of  Analytical
      Methods,  Volume  2, Third Edition, U. S. Department of Health and Human
      Services, February 1984.

10.    Development  of   Methods   for  Sampling 1,3-Butadiene.   Interim  Report
      prepared  under   U.  S.  Environmental Protection Agency Contract Number
      68-02-3993, March 1987.

11.    Hexachlorocyclopentadiene, Method  2518.   NIOSH  Manual  of  Analytical
      Methods,  Volume  2, Third Edition, U. S. Department of Health and Human
      Services, February 1984.

12.    Method 110 -  Determination of  Benzene from  Stationary Sources, Proposed
      Rule.   Federal Register,  Volume 45, No. 77, April 18, 1980, page 26677.

-------
                                                            Section No. 3.16.11
                                                            Date June 30, 1988
                                                            Page 2

13.   Hydrocarbons, Aromatic, Method 1501.  NIOSH Manual of Analytical Methods,
      Volume 2, Third Edition, U. S. Department  of  Health and Human Services,
      February 1984.

14.   Naphthylamines,  Method  264.  NIOSH Manual of Analytical Methods, Volume
      4, Second  Edition, U. S. Department of Health and Human Services, August
    :.  1978.

15.   Ketones I, Method 1300.  NIOSH Manual of  Analytical  Methods,  Volume 2,
      Third  Edition,  U. S. Department of Health and Human Services,  February
      1984.

16.   2-Butanone, Method  2500.   NIOSH Manual of Analytical Methods, Volume 1,
      Third  Edition,  U. S. Department of Health and Human Services,  February
      1984.

17..-  Ethylene Oxide, Method 1607.  NIOSH Manual of Analytical  Methods, Volume
      1, Third Edition, U. S. Department of Health and Human Services, February
      1984.

18.   Propylene Oxide, Method 1612.  NIOSH Manual of Analytical Methods, Volume
      2, Third Edition, U. S. Department of Health and Human Services, February
      1984.

19.   Hydrocarbons, Halogenated,  Method  1003.   NIOSH  Manual  of  Analytical
      Methods, Volume  2,  Third  Edition, U. S. Department of Health and Human
      Services, February 1984.

20.   Ethylene  Dibromide,  Method  1008.   NIOSH Manual of Analytical Methods,
      Volume  1,  Third Edition, U. S.  Department of Health and Human Services,
      February 1984.

21.   Method 23 - Determination of Halogenated Organics from Stationary Sources
      (proposed  method).  Federal Register,  Volume 45, No. 114, June 11, 1980,
      page 39766.

22.   1,2-Dichloropropane,  Method  1013.   NIOSH Manual of Analytical Methods,
      Volume  1,  Third Edition, U. S.  Department of Health and Human Services,
      February 1984.

23.   Development  of Methods for Sampling Chloroform and Carbon Tetrachloride.
      Interim Report prepared for U. S. Environmental  Protection  Agency under
      EPA Contract Number 68-02-3993, November 1986.

24.   Dichlorodifluoromethane, Method 111.  NIOSH Manual of Analytical Methods,
      Volume  2, Second Edition, U. S.  Department of Health and Human Services,
      April 1977-

-------
                                                            Section No.  3.16.11
                                                            Date June 30,  1988
                                                            Page 3


25.   Methyl Bromide, Method 2520.  NIOSH Manual of Analytical  Methods, Volume
      2, Third Edition, U. S. Department of Health and Human Services, February
      1984.

26.   Methyl  Chloride,  Method 99.  NIOSH Manual of Analytical Methods, Volume
      4, Second  Edition, U. S. Department of Health and Human Services, August
      1978.

27.   Butler, F. E., E. A. Coppedge, J. C. Suggs, J. E.  Knoll,  M. R. Midgett,
      A. L.  Sykes, M. W. Hartman, and J.  L.  Steger.   Development of a Method
      for Determination  of Methylene Chloride Emissions at Stationary Sources.
      Paper  for  presentation at 80th Annual Meeting of Air Pollution  Control
      Association, New York, NY, June 1987.

28.   Vinylidene Chloride, Method 266.  NIOSH  Manual  of  Analytical  Methods,
      Volume  4, Second Edition, U. S. Department of Health and Human Services,
      August 1978.

29.   Ethyl Chloride,  Method 2519.  NIOSH Manual of Analytical Methods, Volume
      1, Third Edition, U. S. Department of Health and Human Services, February
      1984.

30.   Method  106  - Determination of Vinyl Chloride from  Stationary  Sources.
      Federal Register, Volume 47, No. 173, September 7, 1982, page 39168.

31.   Knoll,  J.  E.,  M.  A. Smith, and M. R. Midgett.  Evaluation of Emission
      Test Methods for Halogenated Hydrocarbons, Volume II, U. S. Environmental
      Protection Agency Publication No. EPA-600/4- 80-003, January 1980.

32.   Methylene  Chloride, Method 1005.  NIOSH Manual  of  Analytical  Methods,
      Volume 2, Third Edition, U. S. Department  of  Health and Human Services,
      February 1984.

33.   Tetrachloroethylene,  Method  335.   NIOSH  Manual of Analytical Methods,
      Volume 3, Second Edition, U. S. Department  of Health and Human Services,
      April 1977-

34.   Trichloroethylene,  Method  336.   NIOSH  Manual  of  Analytical Methods,
      Volume 3, Second Edition, U. S. Department  of Health and Human Services,
      April 1977-

35.   1,1,2-Trichlorotrifluoroethane,  Method  129.  NIOSH Manual of Analytical
      Methods,  Volume  2, Second Edition, U. S. Department of Health and Human
      Services, April 1977.

36.   Vinyl Chloride, Method 1007-  NIOSH Manual of Analytical  Methods, Volume
      2, Third Edition, U. S. Department of Health and Human Services, February
      1984.

37-   Mann,  J. B., J. J. Freal,  H.  F. Enos, and J. X. Danauskas.  Development
      and  Application  of   Methodology   for   Determining   1,2   Dibromo-3-
      Chloropropane  (DBCP) in Ambient Air.  Journal  of  Environmental  Science
      and Health, B15(5), 519-528 (1980).

-------
                                                             Section No. 3.16.11
                                                             Date June 30, 19S8
                                                             Page 4


38.  VOC  Sampling  and Analysis Workshop,  Volume  III.   U.  S.  Environmental
     Protection Agency Publication No. EPA-340/1-84-001C, September 1984.

39.  Knoll, J. E.,  M. A. Smith, and M. R. Midgett.  Evaluation of Emission Test
     Methods  for  Halogenated  Hydrocarbons,  Volume  I.   U.  S. Environmental
     Protection Agency Publication No. EPA-600/4-79-025, March 1979.

40.  Binetti, R. et al.  Headspace Gas  Chromatographic  Detection  of  Ethylene
     Oxide in Air.  Chromatographia, Vol. 21, December 1986.

4l.  Butadiene,  Method  591.   NIOSH  Manual of Analytical Methods,  Volume  2,
     Second Edition, U. S. Department of Health and Human Services, April 1977-

42.  Knoll,  J.  E.  Estimation of the Limit  of  Detection  in  Chromatography.
     Journal of Chromatographic Science, Vol. 23, September 1985.

43.  Procedure  1 - Determination of Adequate Chromatographic  Peak  Resolution.
     Code of Federal Regulations, Title 40. Part 6l, Appendix C, July 1, 1987.

44.  Method 625  -  Base/Neutrals and Acids.  Code of Federal Regulations, Title
     40, Part 136, Appendix A, July 1, 1987.

45.  C. through C,_ Hydrocarbons in the Atmosphere by Gas Chromatography.  ASTM D
     2Q20-72,  Part  23.   American   Society   for   Testing   and   Materials,
     Philadelphia, PA, 23:950-958, 1973.

46.  Corazon,  V.  V.   Methodology for Collecting  and  Analyzing  Organic  Air
     Pollutants.    U.   S.  Environmental  Protection  Agency  Publication  No.
     EPA-600/2-79-042, February 1979.

47.  Dravnieks,  A.,  B.  K.  Krotoszynski, J. Whitfield, A. O'Donnell,  and  T.
     Burgwald.  Environmental Science and Technology,  5(12):1200-1222, 1971.

48.  Eggertsen, F. T., and F. M. Nelson.  Gas Chromatographic Analysis of Engine
     Exhaust and Atmosphere.  Analytical Chemistry, 30(6):  1040-1043,  1958.

49.  Feairheller,  W. R., P. J. Marn, D. H. Harris, and D.  L.  Harris.   Technical
     Manual  for  Process  Sampling  Strategies for Organic  Materials,   U.   S.
     Environmental Protection Agency, Publication  No.   EPA  600/2-76-122, April
     1976.

50.  FR, 39 FR 9319-9323, 1974.

51.  FR, 39 FR 32857-32860, 1974.

52.  FR, 41 FR 23069-23072 and 23076-23090, 1976.

53.  FR, 41 FR 46569-46571, 1976.

54.  FR, 42 FR 41771-41776, 1977-

-------
                                                             Section No.  3.16.11
                                                             Date June 30,  1988
                                                             Page 5

55.  Fishbein, L.  Chromatography  of Environmental Hazards, Volume II. Elsevier
     Scientific Publishing Company, New York, New York, 1973-

56.  Hamersma, J. W.,  S. L. Reynolds, and R. F. Maddalone.   EPA/IERL Procedures
     Manual:  Level  1 Environmental Assessment, U. S. Environmental  Protection
     Agency Publication No. EPA 600/276/l60a, June 1976.

57.  Harris,  J.  C.,   M.  J.  Hayes, P. L. Levins, and D. B. Lindsay.  EPA/IERL
     Procedures for Level  2  Sampling and Analysis of Organic Materials.  U. S.
     Environmental Protection Agency Publication No. EPA 600/7-79-033,  February
     1979.

58.  Harris, W. E.,  H.  W. Habgood.  Programmed Temperature Gas Chromatography.
     John Wiley & Sons, Inc.  New York,  1966.

59.  Methods  of Air Sampling and Analysis.   Intersociety  Committee,  American
     Health Association, Washington, D. C., 1972.

60.  Jones,  P.  W.,  R.   D.   Grammer,  P.  E.  Strup,  and  T.  B.  Stanford.
     Environmental Science and Technology, 10:806-810, 1976.

6l.  McNair Han Bunelli, E. J. Basic Gas Chromatography.  Consolidated Printers,
     Berkeley, 1969.

62.  Nelson, G. 0. Controlled Test Atmospheres, Principles and Techniques.    Ann
     Arbor, Ann Arbor Science Publishers, 1971-

63.  Schuetzle, D., T. J. Prater, and S. R. Ruddell.  Sampling and  Analysis  of
     Emissions  from  Stationary  Sources;  I.   Odor  and  Total  Hydrocarbons.
     Journal of the Air Pollution Control Association, 1975-

64.  Snyder,  A.  D.,  F. N. Hodgson, M. A. Kemmer, and J. R. McKendree.  Utility
     of Solid Sorbents  for  Sampling Organic Emissions from Stationary Sources.
     U. S. Environmental  Protection  Agency  Publication  No. EPA 600/2-76-201,
     July 1976.

65.  Tentative  Method  for  Continuous  Analysis of Total Hydrocarbons  in  the
     Atmosphere.  Intersociety  Committee,  American  Public Health Association,
     Washington, D.C., 1972.

66.  Zwerg, G.  CRC Handbook of Chromatography, Volumes  I  and  II.  CRC Press,
     Cleveland, 1972.

-------
                                                                Section No. 3.16.12
                                                                Date June 30, 1988
                                                                Page 1
12.0   DATA FORMS

      Blank data forms  are  provided on the following pages  for  the convenience of
 the Handbook user.   Each  blank  form has  the customary descriptive title centered at
 the top of the page.   However,  the  section-page documentation in the top right-hand
 corner of each page has been replaced with a number in the lower right-hand corner
 that will enable the  user  to  identify and refer to a  similar  filled-in  form in a
 text section.  For example, form M18-2.5  indicates  that  the form  is Figure 2.5 in
 Section 3.18.2 of the Method 18 section.  Future revisions of these forms, if any,
 can be documented by 2.5a,  2.5b, etc.  Nineteen of the blank forms listed below are
 included in this  section.  Six have been left blank in the text as shown following
 the form number.
 Form

 2.1A & B


 2.2A & B


 2.5

 2.6

 2.7

 3.1 (Text)

 3.2 (Text)

 3.4 (Text)

 3.5 (Text)

 4.1

 4.2

 4.3

 4.8 (Text)

 5-1

 5-4


 5-6
      Title

Flowmeter Calibration Data Form
(English and metric units)

Critical Orifice Calibration Data Form
(English and metric units)

Dynamic Dilution Data Form

Static Dilution Data Form

Thermometer Calibration Form

Preliminary Survey Data Sheet

Preliminary Survey Preparations

Pretest Sampling Checks

Pretest Preparations

Field Sampling Data Form for Container Sampling

Field Sampling Data Form for Direct Interface Sampling

Field Sampling Data Form for Adsorption Tube Sampling

On-site Measurements Checklist

Data Form for Analysis of Method 18 Samples

Calibration Standard Preparation Data Form for Diluted
Gas Cylinders

Calibration Data Form for Preparation of Standards in
Tedlar Bags by Gas and Liquid Injection

-------
                                                                Section No. 3-16.12
                                                                Date June 30, 1988
                                                                Page 2
5.8                    Data Form for Development of Response and Relative
                       Retention Factors

5.9                    Data Form for Preparation of Liquid Standards and
                       Desorption Efficiency Samples for Adsorption Tube Analysis
5.10 (Text)            Postsampling Operations Checklist

6.1                    Calculation Form for GC Analysis by Gas Injection

6.2                    Calculation Form for GC Analysis by Liquid Injection

8.1                    Field Audit Report Form

8.2                    Method 18 Checklist to be Used by Auditors

-------
                 FLOWMETER CALIBRATION DATA FORM  (English units)
Date
Calibrated by
Meter system no.
Primary meter no.
        op
Barometric pressure, Pn = 	 in. Hg   Ambient temperature 	
Type of primary meter:  wet test 	, dry gas 	, or bubble meter
Type of flowmeter calibrated:  rotameter
                             , dry gas meter
                        ,  or mass flowmeter
Primary meter readings
Initial
reading
(V } a
vvpi ' •
ft3



Final
reading
(vpf),a
ft3



Initial
temp , ° F
(tpl)
oF



Final
temp,°F
(tpf)
op



Press
drop
(DP)C
in.
H20



Flowmeter readings
Initial
reading
(V ),»
ft3 or
ft3 /min



Final
reading
(V"
ft3 or
ft3 /min



Initial
temp
(tsi)
op



Final
temp
(tsf)
op



Press
drop
(Ds),c
in.
H20



Time
min
(e)_d
min



Calibration
factors
(YJ,'



(Yj



a  Volume passing through the meter using the initial and final readings; requires a minimum of at
   least five revolutions of the meter.
b  Volume passing through the meter using the initial and final readings or the indicated flow rate
   using the initial and final flow rate setting.
c  Pressure drop through the meter used to calculate the meter pressure.
d  The time it takes to complete the calibration run.
e  With Y defined as the average ratio of volumes for the primary meter compared to the flowmeter
   calibrated, Yj = Y +_ 0.03Y for the calibration and Yf = Y ^ 0.05Y for the posttest checks; thus:
 Y,  =
   For calibration of the dry gas meter:

                        + tsf)/2 + 460°F][Pm +  (Dp/13.6)]
              -(Eq.2-l),Y =
                                                                          -(Eq.2-2)
      [<*.! + tsf)/2 + 160°F][Pm +  (DpA3.6)]
     [(Vsf * Vs.)/2]0[(t   + t  )/2 + 460°F][Pra *  (D.A3-6)]
                                         -(Eq. 2-3),Y =
                                                                                             -(Eq.2-4)
                                                                      Quality Assurance Handbook M18-2.1A

-------
                              FLOWMETER CALIBRATION DATA FORM (metric units)
Date
Calibrated by
                                          Meter system no.
Barometric pressure, Pm = 	
Type of primary meter: wet test 	•
Type of flowmeter calibrated: rotameter
                       mm Hg  Ambient temperature
Primary meter no.
        °C
                        , dry gas
                                                               ,  or bubble meter
                            ,  dry gas meter
                                                                 ,  or mass flowmeter
Primary meter readings
Initial
reading
(vpl),a
m3



Final
reading.
(vpf),a
m3



Initial
temp,°F
(tpl)
°C



Final
temp,°F
(tpf)
°C



Pres
drop
(Dp)c
mm
H20



Flowmeter readings
Initial
reading
(v?i),b
m* or
m3/min



Final
reading

H20



Time
min
(e)_d
min



Calibration
factors
(YJ,6



(Y)



8  Volume passing through  the meter using  the initial and final readings and requires a minimum of  at
   least five revolutions  of the meter.
b  Volume passing through  the meter using  the initial and final readings or the indicated flow rate
   using the initial and final  flow rate setting.
c  Pressure drop through the meter used to calculate the meter pressure.
d . The time it takes to complete the  calibration run.
e  With Y defined as the average ratio of  volumes for the primary meter compared to the flowmeter
   calibrated, Y. = Y  + 0.03Y for the calibration and Yt = Y + 0.05Y for the posttest checks; thus,

   For calibration of  the  dry gas meter:                 -   .

                     .  + tsf)/2  + 273°K][Pm +  (Dp/13.6)]
                                                          .(Eq. 2-5) •, Y =
                             K 273°K][Pm +- I

For calibration of the rotameter and mass .flowmeter:

  
-------
                            CRITICAL ORIFICE CALIBRATION DATA  FORM  (English units)
Date
    Calibrated by
                                          Meter system no.
Barometric pressure, Pm = 	
Type of primary meter: wet test 	
Type of critical orifice: capillary glass
                         in. Hg  Ambient temperature
                        	,  dry gas 	
Primary meter no.
       op
                                      needle or tubing
                                                               ,  or bubble meter
                                                                          ,  or adjustable
Primary meter readings
Initial
reading
(vpi),a
ft3



Final
reading
(V ) a
*vpf ' •
ft3



Initial
temp , ° F
(tpi)
op



Final
temp , ° F
(tpf)
op



Pres
drop
<°P>
in.
H20



Critical orifice readings
Initial
setting
b
ft3 or
ft3 /min



Final
setting
b
ft3 or
ft3 /min



Press
drop
c
in.
Hg



Time
min
(e) d
min



Calculated
flow rate
[Q(std)]e
ft3 /min



Calibration
factor*
(K'J



(K')



a  Volume passing through the meter using the initial and final readings and requires a minimum of at
   least five revolutions of the meter.
b  Volume passing through the orifice using the initial and final readings or the indicated flow rate
   using the initial and final flow rate setting (for variable setting orifice only).
0  Pressure drop through the meter used to calculate the meter pressure.
d  The time it takes to complete the calibration run.
e  With K1 defined as the average orifice calibration factor based on the volumes of the primary test
   meter, K'A = K' + 0.03K' for the calibration and K'j = K1 + 0.05K1 for the posttest checks: thus,

   Flow rate of the primary meter at standard conditions:
   Vp
     ( s td )
l7.7Kvpf - vpi)(Pm * Dp/13.6)

   [
-------
                         CRITICAL ORIFICE CALIBRATION DATA FORM  (metric units)
Date
Calibrated by
                                          Meter system no.
Barometric pressure, Pm = 	
Type of primary meter: wet test 	
Type of critical orifice: capillary glass
Primary meter no.
        °C
                     mm Hg    Ambient temperature    '	
                     	, dry gas 	, or bubble meter
                                  needle or tubing
                                                                          ,  or adjustable
Primary meter readings
Initial
reading
(vpi).a
L



Final
reading
(V ) a
i%f ' •
L



Initial
temp , ° F
(tpt)
°C



Final
temp,°F
(tpf)
°C



Pres
drop

-------
                           DYNAMIC CALIBRATION DATA FORM
Date    •	
Source flowmeter number 	
Stage 1 flowmeter number 	
Stage 2 flowmeter number 	
Barometric press 	 mm (in.) Hg
Organic compound 	
                       Calibrated by    ,     	
                       Date source metee calibrated
                       Date stage 1 meter calibrated
                       Date stage 2 meter calibrated
                       Heated box temperature 	
                       Leak check for total
system
Certified concentration
              ppmv(X)  Date of calibration curve
STAGE 1
Emission gas flowmeter reading, ml/min (qcl).
Diluent gas flowmeter reading, ml/min (qdl)
Dilution ratio
Injection time, 2kh
Distance to peak, cm
Chart speed, cm/min
Retention time, min
Attenuation factor
Peak area or units
Peak area X attenuation factor
Measured concentration,8 ppmv
Calculated concentration,6 ppmv (Cs)
Percent difference,"1 %

STAGE 2 (if applicable)
Emission gas flowmeter reading, ml/min (qc2).
Diluent gas flowmeter reading, ml/min (qd2)
Dilution ratio
Injection time, 24h
Distance to peak, cm
Chart speed, cm/min
Retention time, min
Attenuation factor
Peak area or units
Peak area X attenuation factor
Measured concentration,3 ppmv
Calculated concentration,d ppmv
Percent difference,0 %
                               RUN 1
RUN 2
                               RUN 1
RUN2
RUN3
  See Figure 5-1 for calculation.
  C.  =
       106 x (X x qj
     = Calculated concentration for single stage

      Calculated Concentration - Measured Concentration
  Percent Difference =
                                                        x 1002
                                    Measured Concentration
  C,  = 106  x X
(q
                   cl
                        qdl)
                                = Calculated cone,  for two stage.
                                                 Quality Assurance Handbook M18-2.5

-------
                             STATIC DILUTION DATA FORM

Date 	     Calibrated by 	
Source flowmeter number 	     Date source meter calibrated _
Dry gas meter number  	     Date dry gas meter calibrated
Ambient temperature 	 °C (°F)     Dry gas meter calib factor  (Y)
Barometric press 	 mm (in.) Hg     Leak check for total system __
Organic compound 	     Vacuum during leak check  	
Certified concen, (X) 	 ppmv     Date of calibration curve 	
                                                RUN 1        RUN 2
Initial dry gas meter reading, L (ft3)      	
Final dry gas meter reading, L (ft3)        	
Volume of diluent gas metered, L (ft3)      	
Gas metered X calibration factor (Y),{V2}   	
Flowmeter sampling rate, ml/min (cfm)       	
Sampling time, min                          	
Sampling rate X sample time, L (ft3),{V1}   	
Dilution ratio                              	
Injection time, 24h                         	
Distance to peak, cm                        	
Chart speed, cm/min                         	
Retention time, min                         	
Attenuation factor                          	
Peak area or units                          	
Peak area X attenuation factor              	
Measured concentration,3 ppmv               	'	
Calculated concentration,13 ppmv,  {Cs}	
Percent difference,0  %                      	  	   	
a  See Figure 5•1 for calculations.        _
                                         X (VJ
b  Calculated concentration (C )  = 	 = 	  ppmv
                             Measured concent - Calculated concent
   Percent difference,  %d =  	;	 X 100 =
                                     Measured concentration

   The percent difference must be less than 10 % absolute.
                                                 Quality Assurance Handbook M18-2.6

-------
                                  THERMOMETER CALIBRATION FORM
Date








Reference
thermometer
type








Calibr
thermc
type








•ated
imeter
use








no.

" •






Ambier
refer8








it temper
calibr?








Measurec
•ature
differ0








values
Boi
refer8








ling wat
calibr"








,er
differ0








Calibrator's
initials








Temperature reading of the reference thermometer in °C or °F.
Temperature reading of the thermometer being calibrated in °C  or °F.
Difference between the reference thermometer and the calibrated thermometer.   This difference must
be less than 3°C (5.4°F) for than initial calibration and 6°C  (10.4°F)  for the calibration check.
                                                                  Quality Assurance Handbook M18-2.7

-------
                            FIELD SAMPLING DATA FORM FOR CONTAINER SAMPLING
Plant 	
City  	
Operator
Date
                                Flowmeter calib.(Y)
                                Container type: bag
Run number
Stack dia, mm
              (in. )
                syringe
                canister
Container volume, 	
Container number 	
Average ( P) 	
                                     Dilution system:  (dynamic)
                                      emission flowsetting 	
                                      diluent flowsetting 	
                                                             liters
                                                       mm (in.)  H2 0
Sample box number
Pitot tube (Cp) _
Static press
                  mm (in.) H20
Initial flowmeter setting
Average stack temp 	
Barometric press 	
                        mm
 °C  (°F)
'(in.) Hg
Dilution system: (static)
 emission flowsetting 	
Final leak check 	 m3/min (cfm)
Vacuum during leak check 	
	 mm (in.) H20
Sampling point location 	
Sampling
time,
min






Total
Clock
time,
24 h







Velocity head
mm ( in . ) H2 0 ,
( P)






Avg
Flowmeter
setting
L/min (ft3 /min)






Avg
stack
°C (°F)






Avg
probe
°C (°F)






Avg
Temperature
sample line
°C (°F)






Avg
readings
flowmeter box
°C (°F)






Avg
container
°C (°F)






Avg
                                                                   Quality Assurance Handbook M18-4.1

-------
                         FIELD  SAMPLING  DATA  FORM  FOR  DIRECT  INTERFACE SAMPLING
Plant 	
City  	
Operator
Date
                     Barometric press 	
                     Initial probe setting
                     Sampling rate 	
                             mm
Hg
                           (in.)
                        _ °C (°F)
                        L/min (cfm)
Run number
Stack dia,
                     Sampling point location
mm (in.)
Meter box number
Stack temp 	
Static press 	
       mm (in.)
H20
Dilution system:
 source flow rate
 diluent flow rate
 diluent flow rate
Dilution ratio
                                             L/min (cfm)
                                             L/min (cfm)
                                             L/min (cfm)
Sample loop volume
Sample loop temp
Column temperature:
initial /
program rate /
final /
Carrier gas flow
ml
Ci O T?\
\ " 1
°C/min
°C/min
°C/min
ml/min
Dilution system check
Final leak check
Vacuum @ check
mm (in.)
                                  H,0
Time of
injection
24 h







Injection
number







Flov
source
ml/min







raieter(s) s
diluent
ml/min







>ettings
diluent
ml/min
1)






stack
°C (°F)







Temporal
probe
°C (°F)







;ure readings
sample line
°C (°F)







injection port
°C (°F)







                                                                     Quality Assurance Handbook M18-4.2

-------
                         FIELD SAMPLING DATA FORM FOR ADSORPTION TUBE SAMPLING
Plant 	
City  	
Operator
Date
Run number
Stack dia,
                     Flowmeter calib.(Y)  _
                     Adsorption tube type:
                      charcoal tube  	
                      silica gel 	
                      other
mm (in.)
Meter box number
Pitot tube (Cp)
Static press 	
       mm (in.)
H20
Adsorption tube number 	
Average ( P) 	 mm
Initial flowmeter setting
Average stack temp 	
Barometric
                                                          (in.)
                                                     H20
           press
                        mm
 °C  (
"(in.)
                                          Dilution system:  (dynamic)
                                           emission flowsetting 	
                                           diluent flowsetting 	
                                          Dilution system:  (static)
                                           emission flowsetting 	
                                          Final leak check 	 m3/min (cfm)
                                          Vacuum during leak check 	
                                          	 mm (in.)
                                          Sampling point location 	
                                                                                         H20
Hg
Sampling
time,
min






Total
Clock
time,
24 h







Velocity head
mm ( in . ) H2 0 ,
( P)






Avg
Flowmeter
setting
L/min (ft3 /min)






Avg
stack
°C (°F)






Avg
Temperature i
probe , line
°C (°F)






Avg
•eadings
adsorp . tube
°C (°F)






Avg
meter
°C (°F)






Avg
Vacuum
mm ( in . ) Hg






Avg
                                                                     Quality Assurance Handbook M18-4.3

-------
                        ANALYSIS OF METHOD 18 FIELD SAMPLES
Date: 	
Location:
          Analyst:
                                      	  Plant:
                                       Sample Type:
Type of Calibration Standard: 	
Number of Standards: 	  Date Prepared:
                                     Target Compound:
                                             Prepared By:
GC Used: 	
Carrier Gas Used: 	
Column Temperatures, Initial:
Sample Loop Volume:
Detector Temp.: 	
                    Column Used:
                         Carrier Gas Flow Rate:
                         	  Program Rate: 	
                 Loop Temperature:
                                                       	  Final:
                                                  Inject. Port Temp.:
             Auxiliary Gases:
Calibration Data
 First analysis/second analysis
  Standard concentration (Cact)
  Flow rate through loop (ml/min)
  Liquid injection volume (tubes)
  Injection time (24-hr clock)
  Chart speed (cm/min)
  Detector attenuation
  Peak retention time (min)
  Peak retention time range (min)
  Peak area
  Peak area x attenuation factor
  Average peak area value (Y)
  Percent deviation from average
  Calculated concentration (Cstd)
  % deviation from actual
                                   Standard 1
                                          Standard 2
                                                                 Standard 3
  Linear regression equation; slope  (m) :
                                          y-intercept  (b) :
Sample Analysis Data                Sample 1
 First analysis/second analysis
  Sample identification            	
  Interface dilution factor        	
  Flow rate through loop (ml/min)      /	
  Liquid injection volume (tubes)      /	
  Injection time (24-hr clock)         /	
  Chart speed (cm/min)                 /	
  Detector attenuation                 /	
  Peak retention time (min)             /	
  Peak retention time range (min)  	
  Peak area                            /	
  Peak area x atten. factor (At/A2)    /	
  Average peak area value (Y)       	
  % deviation from average (%Vav  ) 	
  Calculated concentration (C  )    	
                                                   Sample 2
                                                          Sample 3
             (Y - b)
C
C.  =
                m
                                                       act
                                                 Quality Assurance Handbook M18-5-1

-------
           PREPARATION OF STANDARDS  BY DILUTION OF GAS  CYLINDER STANDARDS
Date:
             Preparer:
Purpose:
Cylinder Component:	
Component Concentration (X-) :
                              	  Source:
                               ppm   Certification Date:
          Stage 1                       Mixture 1      Mixture 2      Mixture 3
     Standard gas flowmeter reading     	      	•	
     Diluent gas flowmeter reading      	      	      	
     Laboratory temperature (°K)	      	      	.
     Barometric pressure (Pb)  (mm Hg)   	      	   •   	
     Flow rate  of cylinder gas (qcl) at
       standard conditions (ml/min)     	      	          •	
     Flow rate of diluent gas   (qdl) at '
       standard conditions (ml/min)
Calculated

Stage
concentration (Cs )
«
2 (if used)


Xxqcl
3d + 3dl
Mixture 1



Mixture 2 Mixture 3
Standard gas flowmeter reading
  from stage 1
Diluent gas flowmeter reading
  into stage 2
Average differential pressure (Pd )
  between stage 1 and 2 (mm H20)
Flow rate of diluted gas
  tec2 actual) at standard
  conditions to stage 2 (ml/min)
Flow rate of diluted gas
             at corrected
                      to
  stage 2 (ml/min)
Flow rate of diluent gas (qd2) at
  standard conditions to
  stage 2 (ml/min)
Calculated concentration (C )
         c2 corr)
       standard conditions
                                   = 3
                                      c 2 actual
                  Cs  =  X x
                                                : 2  c o r r
                             (3d  + 3dl)    (q
                                             c 2  c o r r
                                                 Quality Assurance Handbook M18-5-4

-------
        PREPARATION OF STANDARDS IN TEDLAR BAGS BY GAS AND LIQUID INJECTION
Date: 	  Preparer: 	  Purpose:
Organic Compound:	•      Gas: 	 or Liquid:
Compound Source: 	 Compound Purity (P): 	% Compound Mole Weight (M):
        Gas Injection                   Mixture 1      Mixture 2      Mixture 3
  Bag number or identification          	      	      	
  Dry gas meter calibration factor (y)  	      	      	
  Final gas meter reading, liters       	      	      	
  Initial gas meter reading, liters     	      	      	
  Volume metered (Vm), liters           	      	      	
  Ambient temperature, °C               	      		
  Average gas meter temperature, °C     	      	      	
  Absolute gas meter temp. (Tm), °K     	      	      	
  Barometric pressure (Pb), mm Hg       	      	      	
  Average gas meter pressure, mm Hg     	      	      	
  Absolute gas meter press. (Pra)t nun Hg 	      	      	
  Gas volume injected (Gv), ml          	      	      	
  Syringe temperature (T8), °K          	      	      	
  Absolute syringe pressure (Ps), mm Hg 	      	      	
  Calculated concentration (Cs)         	      	      	

                    ps * Tn
                  "•                                         vs calc
                    Ts x Pm                       Cs corr = 	  x 100*
  cs =	                                    P
 	vm * y	

      Liquid Injection                  Mixture 1      Mixture 2      Mixture 3
  Bag number or identification          	      	      	
  Dry gas meter calibration factor (y)  	      	      	
  Final gas meter reading-  liters       	      	      	
  Initial gas meter reading, liters    . 	      	      	
  Volume metered (Vm), liters           	      	      	
  Average gas meter temperature, °C     	.      	      	
  Absolute gas meter temp.  (Tm ), °K     		      	
  Barometric pressure (Pb), mm Hg       	      	      	
  Average gas meter pressure, mm Hg     	      	      	
  Absolute gas meter press. (Pm)t mm Hg 	      	      	
  Liquid organic density (p), ug/ml	      	      	
  Liquid volume injected (Lv), ul       	      	      	
  Calculated concentration (Cs)         	      	      	

                   '    Lv * P * Tm                          Cs calc
  Cs  = 6.24 x 10"  x 	              Cs co.rr = 	 x 100*
                     M x Vm x y x Pm                            P


                                                 Quality Assurance Handbook M18-5.6

-------
      DEVELOPMENT OF RELATIVE RESPONSE FACTORS AND RELATIVE RETENTION FACTORS
Date: 	  Preparer:
Target Compound: 	
Surrogate Compound: 	
                                    	  Purpose:
                                     Type of Standard:
                                     Type of Standard:
Target Compound Calibration Data        Standard 1
 First analysis/verify analysis
    Standard concentration              	
    Flow rate through loop (ml/min)         /	
    Liquid injection volume (tubes)         /	
    Injection time (24-hr clock)            /	
    Chart speed (cm/min)                    /	
    Detector attenuation                    /	
    Peak retention time (tRxl/tRxf)         /
    Peak retention time range           	
    Peak area                           	(_	
    Peak area x atten. factor (Yi/Yx)       /	
    Verification analysis conc.(Cx)     	
    Percent deviation from actual       	
    Caculated retention time (rTxf)     	
    Percent deviation from actual       	
Linear regression equation;   slope (m ):  	
                                                       Standard 2
                                                                      Standard 3
                                                   y-intercept (b) :
Surrogate Calibration Data
 First analysis/second analysis
    Standard concentration
    Flow rate through loop (ml/min)
    Liquid injection volume (tubes)
    Injection time (24-hr clock)
    Chart speed (cm/mni)
    Detector attenuation
    Peak retention time (tRsl/tRsf)
    Peak retention time range
    Peak area
    Peak area x attenuation factor
Linear regression equation; slope  (m
                                        Standard 1
                                                       Standard 2     Standard 3
                                               (mf)
                                                            y-intercept (b) :
Nonretained peak retention time (tM1/tMf):
Relative Response Factor (FRv):
                           Rx
                                        Relative Retention Factor (rv/Q):
                                                                    X / 5
         ms
   FRX = —
                      -x/s
                                                c  =
                                     ^Mi
                                                      "std
                                                           x F
                                                              Rx
                                               °Mf
                                                 Quality Assurance Handbook M18-5-8

-------
  DATA FORM FOR PREPARATION OF LIQUID STANDARDS AND DESORPTION EFFICIENCY SAMPLES
Date:
Preparer:
Purpose:
Organic Compound: _
Compound Source: 	
Adsorbent Material:
                                      Gas:
       Compound Purity (P):
                 Batch No:
                     or Liquid:
    % Compound Mole Weight (M):
   	 Desorption Solvent:
  Standards in Solvent                  Mixture 1      Mixture 2      Mixture 3
  Desorption solvent volume (Vs),  ml    	      	      	
  Compound spike amount (V0),  ul        	      	      	
  Organic compound density (p), ug/ul   	      	      	
  Standard concentration (C ), ug/ml    	      	      	
  Standards on Adsorbent
  Adsorbent amount, g
  Compound spike amount (V0),  ul
  Organic compound density (p),  ug/ul
  Desorption solvent volume (Vs),  ml
  Desorption time, min
  Standard concentration (C8), ug/ml
                    Mixture 1   Mixture 2   Mixture 3  Blank
  GC Operating Conditions
  Injection port temperature, °C
  Carrier gas flow rate, ml/min
  Column temperature:
       Initial, °C
       Program rate, °C/min
       Final, °C
  Chromatographic Results
  Injection time,24-hr clock
  Distance to peak, cm
  Chart speed, cm/min
  Retention time, min
  Attenuation factor
  Standards in desorption solvent:
       Peak area (Ac), area counts
  Standards and blank from
    adsorbent material:
       Peak area (As and Ab),
          area counts
                    Mixture 1   Mixture 2   Mixture 3  Blank
  Desorption Efficiency Calculation
  Desorption Efficiency (DE),  %
                      Mixture 1    Mixture 2    Mixture 3
  C.  =
       V0 x p x P
       V0  x 100%
                              As - Ab
                         DE = 	 x
                                                 Quality Assurance Handbook M18-5.9

-------
                CALCULATION FORM FOR GC ANALYSIS BY GAS INJECTION









                               SAMPLE CONCENTRATION




                   ppm,   Pr  =	. _  mm Hg,   Ti  =	. _  °K,




               . _  mm Hg,   Tr  =	. _  °K,   Bws  =  0  .	,




                ,    F/  =	
                     F  K
                      r
           PI Tr (1 - BW8K)
                            =  ___ ppm                     Equation 6-1
If applicable.
                                                Quality Assurance Handbook M18-6.1

-------
                CALCULATION FORM FOR GC ANALYSIS BY LIQUID INJECTION
                  SAMPLE VOLUME, DRY BASIS AT STANDARD CONDITIONS
 amb
                         K.   B*  -  0
                              P    V
                              r    v
     v td d    =  0-3858
"If applicable.
                                  - B
                               DESORPTION EFFICIENCY




                                        B  =
                                                                    Equation 6-2
     DE  =  (Qr - B)/Qa  =  0 . __
                                                                     Equation 6-3
W
B
K'  =  0 .
                                SAMPLE CONCENTRATION




                   ug,   Wb  =  ____ ug,   Bp  =  ____ ug,




                   ug,   Vstd  =  ___ -_L,  DE  =  0. __ ,
     C  =
                 V
                                   =  ____ mg/dscm or ug/dsL   Equation 6-
                  std
                                 CONVERSION TO PPM




C  =   ____  mg/dscm or ug/dsL,   MW  =  ___ .  _  ug/ug-mole,
     Cppm   =  24.055 C (MW)
                                             ppm                      Equation 6-5
                                                 Quality Assurance Handbook M18-6.2

-------
                            FIELD  AUDIT REPORT
Part A. - To be filled out by organization supplying audit cylinders.
          1.  Organization supplying audit sample(s) and shipping'address
                                                         ]
          2.  Audit supervisor, organization, and phone number

          3.  Shipping instructions: Name, Address, Attention
          4.
          5-
          6.
   Guaranteed arrival date for cylinders -
   Planned shipping date for cylinders - 	
   Details on audit cylinders from last analysis

a. Date of last analysis 	
b. Cylinder number. 	 	
c. Cylinder pressure psi.
d. Audit gas (es) /balance gas .
e. Audit gas(es) ppm
f . Cylinder construction 	

Low cone.







High cone




., ...


Part B. - To be filled out by audit supervisor.
          1. Process sampled	  '
             Audit location
2.
3.
4.
5.
             Name of individual audit
             Audit date 	
             Audit Results:




d. Measured concentration, ppm
Injection #1* Injection #2* Average 	
e. Actual audit concentration, ppm
f. Audit accuracy:1
High Cone Cylinder 	
Percent1 accuracy =
Measured Cone. - Actual Cone. ^ 100
Actual Cone.

Low.
cone :
cylinder









High
cone.
cylinder









       1Results of two consecutive  injections  that  meet"the'sample analysis
        criteria of the test method.
                                            Quality  Assurance  Handbook M18-8.1

-------
METHOD 18  AUDIT  CHECKLIST
Yes






























No






























Comments






























Operation
PRESAMPLING PREPARATION
1. Knowledge of process operations
2. Results of pretest audit (+ 10% or other value)
3- Calibration of pertinent equipment, in
particular, dry gas meters and other flowmeters
4. Selection and checkout of equipment for proper
sampling and analytical techniques
BAGS - reactivity, condensation, & retention
ADSORPTION TUBES - adsorption & desorption
efficiency
DILUTION SYSTEM - dilution ratio
GC/COLUMN - adequate resolution
GC/DETECTOR - acceptable accuracy & precision
ON-SITE MEASUREMENTS
5- Results of on-site audit {+ 10% or other value)
6. Sampling system properly assembled
7. Based on pitot tube check, is proportional
sampling required (more than 10% flow change)
8. Dilution system check acceptable (if applicable)
9- Sampling system leak check acceptable
10. Proportional sampling properly conducted
11. Constant rate sampling properly conducted
12. Heater systems maintained at proper temperatures
13. Proper number of samples & sampling time
14 . GC properly calibrated
15. Duplicate injections had acceptable precision <5#
16. Recording of pertinent process conditions during
sample collection, samples properly identified,
and calculations properly conducted
POSTSAMPLING
17. Results of off -site audit (+ 10% or other value)
18. GC properly calibrated
19. Duplicate injections had acceptable precision <5#
20. Adsorption efficiency acceptable, >90# on primary
21. Desorption efficiency accep table, >50# recovery
22. Adequate peak resolution
23. Bags passed reaction check, less than 10% change
24. Bags passed retention check, less than 5% retained
25. Flowmeters recalibration acceptable
26. Temperature sensor recalibration acceptable

COMMENTS
                      Quality Assurance  Handbook M18-8.2





                            ft U.S. GOVERNMENT PRINTING OFFICE: 1989-648-163/87062

-------